Science.gov

Sample records for 03-02-1999 bp950005-4 amplicor

  1. Reproducibility problems with the AMPLICOR PCR Chlamydia trachomatis test.

    PubMed Central

    Peterson, E M; Darrow, V; Blanding, J; Aarnaes, S; de la Maza, L M

    1997-01-01

    In an attempt to use an expanded "gold standard" in an evaluation of an antigen detection test for Chlamydia trachomatis, the AMPLICOR (Roche Diagnostics Systems, Inc., Branchburg, N.J.) PCR Chlamydia trachomatis test and culture were used with 591 sets of cervical specimens. Of the 591 specimens assayed, 35 were retested due to either an equivocal result by the PCR (19 samples) or a discrepancy between the results of culture, PCR, and the antigen detection method. During the repeat testing of the samples with equivocal and discrepant results, all but one interpretation change was due to the PCR result. In addition, upon repeat testing the PCR assay value measured in optical density units varied widely for 13 of these specimens. These 13 specimens were then tested in triplicate by the manufacturer with primers to the chlamydia plasmid and in duplicate with primers to the major outer membrane protein. Only 3 of the 13 specimens gave the same interpretation with these five replicates. In summary, reproducibility problems with the AMPLICOR test should be considered before it is incorporated as part of routine testing or used as an expanded gold standard for chlamydia testing. PMID:9157161

  2. Use of Roche AMPLICOR Mycobacterium tuberculosis PCR in Early Diagnosis of Tuberculous Meningitis

    PubMed Central

    Bonington, Alec; Strang, J. I. George; Klapper, Paul E.; Hood, Steven V.; Rubombora, William; Penny, Miranda; Willers, Rose; Wilkins, Edmund G. L.

    1998-01-01

    Several nucleic acid-based amplification tests are available for the detection of Mycobacterium tuberculosis, but few data are available on their use in the diagnosis of tuberculous meningitis (TBM). We performed a prospective study to assess the Roche AMPLICOR Mycobacterium tuberculosis PCR test (TB AMPLICOR) for use in the diagnosis of TBM and compared it with direct Ziehl-Neelsen staining of smears, radiometric culture for M. tuberculosis, and clinical and cerebrospinal fluid (CSF) findings. Eighty-three CSF specimens collected from 69 patients with suspected meningitis in South Africa were tested by TB AMPLICOR. On the basis of clinical and laboratory findings, 40 of these patients were treated for TBM and 29 patients were not treated for TBM. Ten CSF samples from 10 patients were positive by TB AMPLICOR. Seven of these 10 patients were classified as having definite TBM, 2 were classified as having probable TBM, and 1 was classified as having possible TBM. The sensitivity of TB AMPLICOR for detecting cases of definite and probable TBM in patients from whom CSF specimens had been collected less than 10 days into antituberculosis treatment was 60.0%. Specimens from all 29 patients not treated for TBM were negative by the TB AMPLICOR, giving a 100% specificity. TB AMPLICOR is therefore more sensitive than the combination of Ziehl-Neelsen staining of smears and radiometric culture for M. tuberculosis and is a rapid and highly specific diagnostic test for TBM. PMID:9574686

  3. Comparison of Amplicor and GeneXpert MTB/RIF Tests for Diagnosis of Tuberculous Meningitis

    PubMed Central

    Patel, Vinod B.; Connolly, Cathy; Singh, Ravesh; Lenders, Laura; Matinyenya, Brian; Theron, Grant; Ndung'u, Thumbi

    2014-01-01

    There are no data about the comparative accuracy of commercially available nucleic acid amplification tests (GeneXpert MTB/RIF and Roche Amplicor) for the diagnosis of tuberculous meningitis (TBM). A total of 148 patients with suspected TBM were evaluated, and cultures served as the reference standard. The sensitivities and specificities (95% confidence interval [CI]) for the Amplicor and Xpert MTB/RIF tests were similar: 46 (31–60) versus 50 (33–67) and 99 (93–100) and 94 (84–99), respectively. PMID:25056328

  4. Performance of the Amplicor human immunodeficiency virus type 1 PCR and analysis of specimens with false-negative results.

    PubMed Central

    Barlow, K L; Tosswill, J H; Parry, J V; Clewley, J P

    1997-01-01

    Over a 4-year period, the Roche Amplicor kit was used in a United Kingdom reference laboratory for the detection or confirmation of human immunodeficiency virus (HIV) type 1 infection, particularly in infants born to HIV-infected mothers. Of 408 specimens from adults and older children tested, the 122 seronegative specimens were all Amplicor negative. Of the 286 seropositive specimens, 268 were Amplicor positive. On the basis of these results, the Amplicor assay has a specificity of 100% and a sensitivity of 93.7%. In addition, for 247 specimens from infants and young children, serological results may not have been diagnostic because of placental transfer of maternal antibodies. Forty-eight were Amplicor positive, and of the 199 Amplicor-negative specimens, 19 were assumed to be false negative on the basis of clinical data, serological markers (including p24 antigen), and/or results for previous or follow-up specimens. This represents a sensitivity of 75% for the Amplicor test for specimens from patients under 2 years of age. Of these 37 false-negative specimens plus 2 specimens from other laboratories, 31 could be characterized by amplifying extracted material from them by an in-house nested gag PCR spanning the Amplicor target region. The amplicons were sequenced and found to represent subtypes A (35.5%), B (22.6%), C (22.6%), D (16.1%), and G (3.2%). False-negative results by the Amplicor assay may be ascribed to low-target copy number, the physical behavior of one primer (SK462), and sequence variation in the target region of the other primer (SK431). PMID:9350745

  5. Comparison of AMPLICOR and Hybrid Capture II assays for high risk HPV detection in normal and abnormal liquid-based cytology: use of INNO-LiPA Genotyping assay to screen the discordant results.

    PubMed

    Mo, L Z; Monnier-Benoit, S; Kantelip, B; Petitjean, A; Riethmuller, D; Prétet, J L; Mougin, C

    2008-02-01

    The study was aimed to evaluate the feasibility of detecting human papillomavirus (HPV) in women with normal or abnormal cervical smears using the Roche Amplicor MWP HPV Test. We compared by AMPLICOR Test and Hybrid Capture II (HCII) Test, the prevalence of HR-HPV in 470 cervical samples including 55 samples with WNL cytology, 208 ASC-US, 193 LGSIL and 14 HGSIL. Samples with discordant results were retested with INNO-LiPA Genotyping HPV Test v2. The HR-HPV positivity in WNL cytology samples was similar (21.8%) by AMPLICOR and HCII. In ASC-US, the HPV positivity was 42.3% by both tests. In LGSIL, HPV positivity was 66.3% and 66.8% by AMPLICOR and HCII, respectively. In HGSIL, 92.8% of samples were positive by AMPLICOR and 85.7% by HCII. The agreement of both tests was 96.2% with a Kappa value of 0.92. Eighteen cases were discordant: 9 HCII positive/AMPLICOR negative and 9 HCII negative/AMPLICOR positive. The INNO-LiPA test revealed HPV positivity in every case. Interestingly, all HCII+/AMPLICOR- samples were found to harbour HPV53. As for the HCII-/AMPLICOR+ samples, 8 demonstrated a multiple infection with HR 16- and/or 18- and/or 56-phylogenetically related HPV types. Moreover, two of these samples were co-infected with HPV6 and two other with HPV54. By using consensus HR-HPV as our reference HPV positivity, the sensitivity (96.6%) and specificity (100%) of AMPLICOR was similar to that of HCII Test. The AMPLICOR HPV Test is sensitive, specific, feasible and appropriate for routine HPV detection. PMID:18036888

  6. Cytomegalovirus quantification in plasma with Abbott RealTime CMV and Roche Cobas Amplicor CMV assays.

    PubMed

    Tremblay, Maxime-Antoine; Rodrigue, Marc-André; Deschênes, Louise; Boivin, Guy; Longtin, Jean

    2015-12-01

    We assessed the performance of Abbott RealTime CMV assay (ARC) compared to Roche Cobas Amplicor CMV Monitor Test (RCM) for quantification of CMV in plasma of transplant patients. Commercial panels were used to test linearity, precision and interference and 83 clinical samples were used for the accuracy and precision analyses. All 43 RCM-positive clinical samples tested positive by ARC. The overall concordance between the two tests was good (98%). Based on 17 samples, the inter-assay median coefficient of variation was 13%. A linearity panel ranging from approximately 1 to 7log10copies/mL was used to confirm linearity (R(2)=0.99). CMV viral load measurement was not affected by different concentrations of HSV-1 or EBV DNA. We conclude that The Abbott RealTime CMV assay offers good sensitivity, precision and linearity and is suitable for monitoring CMV viral loads in transplant recipients. Standardization with the WHO CMV standard allows for comparison with other assays. PMID:26341060

  7. Assessment of agreement between the AMPLICOR HIV-1 MONITOR test versions 1.0 and 1.5.

    PubMed

    Hill, Charles E; Green, Alicia M; Ingersoll, Jessica; Easley, Kirk A; Nolte, Frederick S; Caliendo, Angela M

    2004-01-01

    The agreement of the microwell plate AMPLICOR HIV-1 MONITOR version 1.0 (MWP 1.0), the microwell plate AMPLICOR HIV-1 MONITOR version 1.5 (MWP 1.5), and the COBAS AMPLICOR HIV-1 MONITOR version 1.5 (COBAS 1.5) tests was evaluated using clinical specimens and well-characterized control material. Two hundred patient plasma specimens and a panel of known human immunodeficiency virus type 1 (HIV-1) subtypes were tested. All data were log(10) transformed prior to analysis. The 95% limits of agreement for the three tests at the average of 3.66 log(10) copies/ml were +/- 0.28 log(10), +/- 0.34 log(10), and +/- 0.34 log(10) copies/ml for MWP 1.0-MWP 1.5, MWP 1.0-COBAS 1.5, and MWP 1.5-COBAS 1.5, respectively. Ten specimens (6.1%) had differences exceeding the limits of agreement for the MWP 1.0 and MWP 1.5 tests. Correlation coefficients among the three tests were high (r >or=0.96). The viral-load values obtained with the MWP 1.0 test were only 2.1% higher on average than those measured with the MWP 1.5 test and 1.6% higher than those seen with the COBAS 1.5 test. The MWP 1.5 test values were 0.8% higher than the COBAS 1.5 test values. Overall, there was less agreement among the different tests for viral-load values near the lower limit of quantification. The MWP 1.0 test underquantified subtypes A, E, F, G, and H by 1.0 to 2.0 log(10) copies/ml; this problem was not observed with the MWP 1.5 test. The close agreement among the results obtained with the different test versions and formats suggests that it is not necessary to reestablish a baseline viral load when changing AMPLICOR HIV-1 MONITOR tests, unless the patient is known to be infected with a non-B subtype. PMID:14715766

  8. Diagnostic value of the strand displacement amplification method compared to those of Roche Amplicor PCR and culture for detecting mycobacteria in sputum samples.

    PubMed Central

    Ichiyama, S; Ito, Y; Sugiura, F; Iinuma, Y; Yamori, S; Shimojima, M; Hasegawa, Y; Shimokata, K; Nakashima, N

    1997-01-01

    We compared the ability of the semiautomated BDProbeTec-SDA system, which uses the strand displacement amplification (SDA) method, with that of the Roche Amplicor-PCR system and the Septi-Chek AFB culture system to directly detect Mycobacterium tuberculosis complex (MTB) and other mycobacteria in sputum samples. A total of 530 sputum samples from 299 patients were examined in this study. Of the 530 samples, 129 were culture positive for acid-fast bacilli with the Septi-Chek AFB system; 95 for MTB, 29 for M. avium-M. intracellulare complex (MAC), and 5 for other mycobacteria. The BDProbeTec-SDA system detected 90 of the 95 samples culture positive for MTB (sensitivity, 94.7%), and the Amplicor-PCR system detected 85 of the 95 samples culture positive for MTB (sensitivity, 89.5%). The specificity of each system, based on the clinical diagnosis, was 99.8% for SDA and 100% for PCR, respectively. Among the 29 samples culture positive for MAC, the BDProbeTec-SDA system detected MAC in 24 samples (sensitivity, 82.8%), whereas the Amplicor-PCR system detected MAC in 23 samples (sensitivity, 79.3%). The specificities of the systems were 98.3 and 100%, respectively. The high degrees of sensitivity and specificity of the BDProbeTec-SDA system suggest that it should be very useful in clinical laboratories for the rapid detection of mycobacteria in sputum samples. PMID:9399498

  9. Analysis of cytomegalovirus (CMV) viremia using the pp65 antigenemia assay, the amplicor CMV test, and a semi-quantitative polymerase chain reaction test after allogeneic marrow transplantation.

    PubMed

    Ksouri, H; Eljed, H; Greco, A; Lakhal, A; Torjman, L; Abdelkefi, A; Ben Othmen, T; Ladeb, S; Slim, A; Zouari, B; Abdeladhim, A; Ben Hassen, A

    2007-03-01

    A pp65 antigenemia assay for polymorphonuclear leukocytes (PMNLs) (CINAkit Rapid Antigenemia), and a qualitative polymerase chain reaction (PCR) test for plasma 'PCR-P qual' (Amplicor cytomegalovirus [CMV] test) were performed for 126 samples (blood and plasma) obtained from 18 bone marrow transplant patients, over a 9-month surveillance period. Among those samples, 92 were assayed with a semi-quantitative PCR test for PMNLs 'PCR-L quant.' The number of samples with a positive CMV test for antigenemia and PCR-P qual assays was 20.63% and 12.7%, respectively, whereas the PCR-L quant assay was positive in 48 of the 92 samples assayed (52.17%). The rates of concordance of the results of PCR-P qual and antigenemia, PCR-P qual and PCR-L quant, antigenemia and PCR-L quant were 92%, 65.2% and 66.8%, respectively. The analysis of the results for the 92 specimens tested by all 3 methods showed a rate of concordance of 63% among all methods. Good agreement (kappa=0.72) was found only between pp65 Ag and PCR-P qual assays. Clinical disease correlates with an antigenemia high viral load. Three patients had CMV disease despite preemptive therapy, and all of them had graft-versus-host-disease (GVHD). PMNLs-based assays are more efficient in monitoring CMV reactivation, but for high-risk patients with GVHD, more sensitive assays (real-time PCR) must be done. PMID:17313466

  10. Comparison of an in-house PCR assay, direct fluorescence assay and the Roche AMPLICOR Chlamydia trachomatis kit for detection of C. trachomatis.

    PubMed

    Sachdeva, Poonam; Patel, Achchhe Lal; Sachdev, Divya; Ali, Mashook; Mittal, Aruna; Saluja, Daman

    2009-07-01

    To improve the control of Chlamydia trachomatis infection in India, a rapid, specific and cost-effective method is much needed. We developed an in-house PCR assay by targeting a unique genomic sequence encoding a protein from the C. trachomatis phospholipase D endonuclease superfamily that produces an amplified fragment of 368 bp. The specificity of the primers was confirmed using genomic DNA from other sexually transmitted disease-causing and related micro-organisms and from humans. The assay was highly sensitive and could detect as low as 10 fg C. trachomatis DNA. Clinical evaluation of the in-house-developed PCR was carried out using 450 endocervical specimens that were divided in two groups. In group I (n=274), in-house PCR was evaluated against the direct fluorescence assay. The resolved sensitivity of the in-house PCR method was 97.22 % compared with 88 % for the direct fluorescent antibody assay. In group II (n=176), the in-house PCR was compared with the commercial Roche AMPLICOR MWP CT detection kit. The resolved sensitivity of the in-house PCR assay reported here was 93.1 % and the specificity was 97.46 %, making it a cost-effective alternative for routine diagnosis of genital infection by C. trachomatis. The method should facilitate early detection leading to better prevention and treatment of genital infection in India. PMID:19502371

  11. Evaluation of the NucliSens EasyQ v2.0 Assay in Comparison with the Roche Amplicor v1.5 and the Roche CAP/CTM HIV-1 Test v2.0 in Quantification of C-Clade HIV-1 in Plasma

    PubMed Central

    Muenchhoff, Maximilian; Madurai, Savathee; Hempenstall, Allison Jo; Adland, Emily; Carlqvist, Anna; Moonsamy, Angeline; Jaggernath, Manjeetha; Mlotshwa, Busisiwe; Siboto, Emma; Ndung'u, Thumbi; Goulder, Philip Jeremy Renshaw

    2014-01-01

    Human immunodeficiency virus type 1 (HIV-1) genetic diversity poses a challenge to reliable viral load monitoring. Discrepancies between different testing platforms have been observed, especially for non-clade-B virus. Therefore we compare, in antiretroviral therapy (ART)-naïve South African subjects predominantly infected with HIV-1 clade-C, three commercially available assays: the COBAS AmpliPrep/COBAS TaqMan HIV-1 Test version 2.0 by Roche (CAP/CTM v2.0), the BioMérieux NucliSens Version 2.0 Easy Q/Easy Mag (NucliSens v2.0) and the Roche COBAS Amplicor HIV-1 Monitor Test Version 1.5 (Amplicor v1.5). Strong linear correlation was observed and Bland-Altman analyses showed overall good agreement between the assays with mean viral load differences of 0.078 log cp/ml (NucliSens v2.0 – Amplicor v1.5), 0.260 log cp/ml (CAP/CTM v2.0 – Amplicor v1.5) and 0.164 log cp/ml (CAP/CTM v2.0 – NucliSens v2.0), indicating lower mean viral load results for the Amplicor v1.5 and higher mean readings for the CAP/CTM v2.0. Consistent with observations following previous comparisons of CAP/CTM v2.0 versus Amplicor v1.5, the CAP/CTM v2.0 assay detected low-level viremia (median 65 cp/ml) in more than one-third of those in whom viremia had been undetectable (<20 cp/ml) in assays using the NucliSens platform. These levels of viremia are of uncertain clinical significance but may be of importance in early detection of ART resistance in those on treatment. Overall the three assays showed good comparability of results but with consistent, albeit relatively small, discrepancies for HIV-1 clade-C samples, especially in the low-viremic range that should be taken into account when interpreting viral load data. PMID:25157919

  12. An in-house real-time polymerase chain reaction: standardisation and comparison with the Cobas Amplicor HBV monitor and Cobas AmpliPrep/Cobas TaqMan HBV tests for the quantification of hepatitis B virus DNA

    PubMed Central

    Santos, Ana Paula de Torres; Levi, José Eduardo; Lemos, Marcilio Figueiredo; Calux, Samira Julien; Oba, Isabel Takano; Moreira, Regina Célia

    2016-01-01

    This study aimed to standardise an in-house real-time polymerase chain reaction (rtPCR) to allow quantification of hepatitis B virus (HBV) DNA in serum or plasma samples, and to compare this method with two commercial assays, the Cobas Amplicor HBV monitor and the Cobas AmpliPrep/Cobas TaqMan HBV test. Samples from 397 patients from the state of São Paulo were analysed by all three methods. Fifty-two samples were from patients who were human immunodeficiency virus and hepatitis C virus positive, but HBV negative. Genotypes were characterised, and the viral load was measure in each sample. The in-house rtPCR showed an excellent success rate compared with commercial tests; inter-assay and intra-assay coefficients correlated with commercial tests (r = 0.96 and r = 0.913, p < 0.001) and the in-house test showed no genotype-dependent differences in detection and quantification rates. The in-house assay tested in this study could be used for screening and quantifying HBV DNA in order to monitor patients during therapy. PMID:26872342

  13. Comparison of the Digene Hybrid Capture 2 Assay and Roche AMPLICOR and LINEAR ARRAY Human Papillomavirus (HPV) Tests in Detecting High-Risk HPV Genotypes in Specimens from Women with Previous Abnormal Pap Smear Results▿

    PubMed Central

    Stevens, Matthew P.; Garland, Suzanne M.; Rudland, Elice; Tan, Jeffrey; Quinn, Michael A.; Tabrizi, Sepehr N.

    2007-01-01

    The development of cervical cancer is strongly associated with the presence of persistent high-risk (HR) human papillomavirus (HPV) infection. Recently, the commercially manufactured PCR-based Roche AMPLICOR (AMP) and LINEAR ARRAY (LA) HPV tests have become available for HPV detection. However, knowledge of their clinical performance compared to the U.S. Food and Drug Administration-approved Hybrid Capture 2 (HC2) assay is limited. This study evaluated the concordance between the HC2, AMP, and LA tests in detecting HR-HPV among a cohort of 1,679 women with previous abnormal Pap smear results. Overall, 1,393 specimens (81.3%) generated concordant results for HR-HPV presence or absence by the three assays. The concordance levels were substantial between the HC2 and AMP tests (84.4%, κ = 0.6419) and between the HC2 and LA tests (84.0%, κ = 0.6341) and nearly perfect between the AMP and LA tests (97.8%, κ = 0.9441). HR-HPV prevalence, as detected by the AMP or LA tests, was significantly higher among women with cytological or histological high-grade disease (CIN2 or greater) than that detected by HC2 (P < 0.0001). The AMP and LA tests exhibited greater sensitivity, but lower specificity, than HC2 for detecting HR-HPV among this cohort of women with underlying cervical abnormalities, particularly among subjects with histologically proven high-grade disease. Both PCR-based HPV tests may be valuable in the management of care for women with underlying cervical abnormalities, in predicting treatment success, and in studying the clearance or acquisition of new infections. PMID:17494721

  14. Impact of Human Immunodeficiency Virus Type 1 (HIV-1) Genetic Diversity on Performance of Four Commercial Viral Load Assays: LCx HIV RNA Quantitative, AMPLICOR HIV-1 MONITOR v1.5, VERSANT HIV-1 RNA 3.0, and NucliSens HIV-1 QT

    PubMed Central

    Swanson, Priscilla; de Mendoza, Carmen; Joshi, Yagnya; Golden, Alan; Hodinka, Richard L.; Soriano, Vincent; Devare, Sushil G.; Hackett, John

    2005-01-01

    Human immunodeficiency virus type 1 (HIV-1) evolution and changing strain distribution present a challenge to nucleic acid-based assays. Reliable patient monitoring of viral loads requires the detection and accurate quantification of genetically diverse HIV-1. A panel of 97 HIV-1-seropositive plasma samples collected from Cameroon, Brazil, and South Africa was used to compare the performance of four commercially available HIV RNA quantitative tests: Abbott LCx HIV RNA Quantitative assay (LCx), Bayer Versant HIV-1 RNA 3.0 (bDNA), Roche AMPLICOR HIV-1 MONITOR v1.5 (Monitor v1.5), and bioMérieux NucliSens HIV-1 QT (NucliSens). The panel included group M, group O, and recombinant viruses based on sequence analysis of gag p24, pol integrase, and env gp41. The LCx HIV assay quantified viral RNA in 97 (100%) of the samples. In comparison, bDNA, Monitor v1.5, and NucliSens quantified viral RNA in 96.9%, 94.8%, and 88.6% of the samples, respectively. The two group O specimens were quantified only by the LCx HIV assay. Analysis of nucleotide mismatches at the primer/probe binding sites for Monitor v1.5, NucliSens, and LCx assays revealed that performance characteristics reflected differences in the level of genetic conservation within the target regions. PMID:16081923

  15. Benefit of hepatitis C virus core antigen assay in prediction of therapeutic response to interferon and ribavirin combination therapy.

    PubMed

    Takahashi, Masahiko; Saito, Hidetsugu; Higashimoto, Makiko; Atsukawa, Kazuhiro; Ishii, Hiromasa

    2005-01-01

    A highly sensitive second-generation hepatitis C virus (HCV) core antigen assay has recently been developed. We compared viral disappearance and first-phase kinetics between commercially available core antigen (Ag) assays, Lumipulse Ortho HCV Ag (Lumipulse-Ag), and a quantitative HCV RNA PCR assay, Cobas Amplicor HCV Monitor test, version 2 (Amplicor M), to estimate the predictive benefit of a sustained viral response (SVR) and non-SVR in 44 genotype 1b patients treated with interferon (IFN) and ribavirin. HCV core Ag negativity could predict SVR on day 1 (sensitivity = 100%, specificity = 85.0%, accuracy = 86.4%), whereas RNA negativity could predict SVR on day 7 (sensitivity = 100%, specificity = 87.2%, accuracy = 88.6%). None of the patients who had detectable serum core Ag or RNA on day 14 achieved SVR (specificity = 100%). The predictive accuracy on day 14 was higher by RNA negativity (93.2%) than that by core Ag negativity (75.0%). The combined predictive criterion of both viral load decline during the first 24 h and basal viral load was also predictive for SVR; the sensitivities of Lumipulse-Ag and Amplicor-M were 45.5 and 47.6%, respectively, and the specificity was 100%. Amplicor-M had better predictive accuracy than Lumipulse-Ag in 2-week disappearance tests because it had better sensitivity. On the other hand, estimates of kinetic parameters were similar regardless of the detection method. Although the correlations between Lumipulse-Ag and Amplicor-M were good both before and 24 h after IFN administration, HCV core Ag seemed to be relatively lower 24 h after IFN administration than before administration. Lumipulse-Ag seems to be useful for detecting the HCV concentration during IFN therapy; however, we still need to understand the characteristics of the assay. PMID:15634970

  16. EUROarray human papillomavirus (HPV) assay is highly concordant with other commercial assays for detection of high-risk HPV genotypes in women with high grade cervical abnormalities.

    PubMed

    Cornall, A M; Poljak, M; Garland, S M; Phillips, S; Machalek, D A; Tan, J H; Quinn, M A; Tabrizi, S N

    2016-06-01

    The purpose of this study was to evaluate the performance of the EUROIMMUN EUROArray HPV genotyping assay against the Roche Cobas 4800, Roche HPV Amplicor, Roche Linear Array and Qiagen Hybrid Capture 2 assays in the detection of high-risk HPV (HR-HPV) from liquid based cervical cytology samples collected from women undergoing follow-up for abnormal cervical cytology results. Cervical specimens from 404 women undergoing management of high-grade cytological abnormality were evaluated by EUROarray HPV for detection of HR-HPV genotypes and prediction of histologically-confirmed cervical intraepithelial neoplasia grade 2 or higher (≥CIN2). The results were compared to Hybrid Capture 2, Cobas 4800 HPV, Amplicor and Linear Array HPV. Positivity for 14 HR-HPV types was 80.0 % for EUROarray (95 % CI; 75.7-83.8 %). Agreement (κ, 95 % CI) between the EUROarray and other HPV tests for detection of HR-HPV was good to very good [Hybrid Capture κ = 0.62 (0.54-0.71); Cobas κ = 0.81 (0.74-0.88); Amplicor κ = 0.68 (0.60-0.77); Linear Array κ = 0.77 (0.70-0.85)]. For detection of HR-HPV, agreement with EUROarray was 87.90 % (Hybrid Capture), 93.58 % (Cobas), 92.84 % (Amplicor) and 92.59 % (Linear Array). Detection of HR-HPV was not significantly different between EUROarray and any other test (p < 0.001). EUROarray was concordant with other assays evaluated for detection of high-risk HPV and showed sensitivity and specificity for detection of ≥ CIN2 of 86 % and 71 %, respectively. PMID:27048314

  17. Performance of Roche CAP/CTM HIV-1 qualitative test version 2.0 using dried blood spots for early infant diagnosis.

    PubMed

    Gueye, Sokhna Bousso; Diop-Ndiaye, Halimatou; Diallo, Mamadou Malick; Ly, Omar; Sow-Ndoye, Aissatou; Diagne-Gueye, Ndèye Diabou; Kébé-Fall, Khady; Diop, Fatou; Gaye-Diallo, Aïssatou; Belec, Laurent; Mboup, Souleymane; Touré-Kane, Coumba

    2016-03-01

    In the context of early infant diagnosis (EID) decentralization in sub-Saharan Africa, dried blood spot (DBS) is now widely used for HIV proviral DNA detection in resource-limited settings. A new version of CAP/CTM (version 2) has been introduced, recently by Roche Diagnosis as a new real-time PCR assay to replace previous technologies on qualitative detection of HIV-1 DNA using whole blood and DBS samples. The objective of this study was to evaluate CAP/CTM version 2 compared to CAP/CTM version 1 and Amplicor on DBS. A total of 261 DBS were collected from children aged 4 weeks to 17 months born from HIV-seropositive mothers and tested by the three techniques. CAP/CTM version 2 showed 100% of agreement with Amplicor including 74 positive results and 187 negative results. CAP/CTM version 2 versus CAP/CTM version 1 as well as CAP/CTM version 1 versus Amplicor showed two discordant results giving a sensitivity of 98.6%, specificity of 99.5%, positive predictive value of 98.6% and negative predictive value of 99.5%. The concordance was 99.12% (95% of confidence interval) giving a Kappa coefficient of 0.97 (p<0.001). These findings confirmed the expected good performance of CAP/CTM version 2 for HIV-1 EID. PMID:26706730

  18. Chemiluminescence enzyme immunoassay for monitoring hepatitis C virus core protein during interferon-alpha2b and ribavirin therapy in patients with genotype 1 and high viral loads.

    PubMed

    Enomoto, Masaru; Nishiguchi, Shuhei; Tamori, Akihiro; Kohmoto, Modoka; Habu, Daiki; Sakaguchi, Hiroki; Takeda, Tadashi; Kawada, Norifumi; Seki, Shuichi; Shiomi, Susumu

    2005-09-01

    This study evaluated an updated chemiluminescence enzyme immunoassay (CLEIA) for hepatitis C virus (HCV) core protein for monitoring viral kinetics during treatment with interferon (IFN)-alpha and ribavirin. Using the CLEIA, serum levels of HCV core protein were measured in 17 patients with genotype 1 and high baseline viral loads during the first 4 weeks of combination therapy. HCV RNA was measured by the Amplicor Monitor test for comparison. At the start of therapy, the median HCV level (interquartile range) was 700 (540-940) kIU/ml of viral RNA and 11,310 (5,528-14,238) fmol/L of core protein. HCV RNA was above the upper limit of the linear range of the Amplicor Monitor test in 13 of the 17 patients, while the core protein level was within the linear range of the CLEIA in all patients. During therapy, the proportion of patients with HCV levels below the cutoff values at each time point was less with the Amplicor Monitor test than with CLEIA. Serum HCV core protein level decreased rapidly during the first 24 hr of therapy and more slowly thereafter, with median exponential decays of 1.08 and 0.046 log10/day, respectively. In the second phase, between day 1 and 28, the median decrease in HCV core protein level was higher in four patients with sustained virologic response (0.13 log10/day) than in 13 patients with no response (0.028 log10/day, P = 0.042). The wide linear range of the HCV core protein assay is appropriate for measuring viral loads during therapy with IFN-alpha and ribavirin. PMID:16032731

  19. Development and Evaluation of a Next-Generation Digital PCR Diagnostic Assay for Ocular Chlamydia trachomatis Infections

    PubMed Central

    Last, Anna; Molina-Gonzalez, Sandra; Cassama, Eunice; Butcher, Robert; Nabicassa, Meno; McCarthy, Elizabeth; Burr, Sarah E.; Mabey, David C.; Bailey, Robin L.; Holland, Martin J.

    2013-01-01

    Droplet digital PCR (ddPCR) is an emulsion PCR process that performs absolute quantitation of nucleic acids. We developed a ddPCR assay for Chlamydia trachomatis infections and found it to be accurate and precise. Using PCR mixtures containing plasmids engineered to include the PCR target sequences, we were able to quantify with a dynamic range between 0.07 and 3,160 targets/μl (r2 = 0.9927) with >95% confidence. Using 1,509 clinical conjunctival swab samples from a population in which trachoma is endemic in Guinea Bissau, we evaluated the specificity and sensitivity of the quantitative ddPCR assay in diagnosing ocular C. trachomatis infections by comparing the performances of ddPCR and the Roche Amplicor CT/NG test. We defined ddPCR tests as positive when we had ≥95% confidence in a nonzero estimate of target load. The sensitivity of ddPCR against Amplicor was 73.3% (95% confidence interval [CI], 67.9 to 78.7%), and specificity was 99.1% (95% CI, 98.6 to 99.6%). Negative and positive predictive values were 94.6% (95% CI, 93.4 to 95.8%) and 94.5% (95% CI, 91.3 to 97.7%), respectively. Based on Amplicor CT/NG testing, the estimated population prevalence of C. trachomatis ocular infection was ∼17.5%. Receiver-operator curve analysis was used to select critical cutoff values for use in clinical settings in which a balance between higher sensitivity and specificity is required. We concluded that ddPCR is an effective diagnostic technology suitable for both research and clinical use in diagnosing ocular C. trachomatis infections. PMID:23637300

  20. Clinical Evaluation of COBAS TaqMan PCR for the Detection of Mycobacterium tuberculosis and M. avium Complex

    PubMed Central

    Ikegame, Satoshi; Sakoda, Yoritake; Fujino, Nao; Taguchi, Kazuhito; Kawasaki, Masayuki; Kajiki, Akira

    2012-01-01

    A retrospective observational study was performed to determine the sensitivity and limitation of PCR test for the detection of Mycobacterium tuberculosis and M. avium complex. We obtained clinical specimens collected from the respiratory tract, cultured M. tuberculosis or M. avium complex, and performed PCR analysis. A total of 299 samples (M. tuberculosis, 177; M. avium, 35; M. intracellulare, 87) were analyzed by COBAS TaqMan PCR from April 2007 to March 2011. The PCR positivity rates were 50–55%, 70–100%, 88–98%, and 100% in smear-negative, smear 1+, 2+, and 3+ groups, respectively. The PCR positivity of tuberculosis in smear 1+ was 80.6%, which was statistically significantly (P < 0.001) lower than that of smear 2+ (97.3%). From January 2005 to March 2007, we collected an additional 138 samples (M. tuberculosis, 74; M. avium, 21; M. intracellulare, 43), which were analyzed by COBAS Amplicor PCR. The PCR positivity rates obtained using COBAS TaqMan PCR and COBAS Amplicor PCR were not significantly different. The sensitivity of PCR test for mycobacteria is not sufficient in case of smear 1+. Careful consideration must be given to the interpretation of negative PCR test results in smear 1+, because smear-positive tuberculosis is the criterion for isolation. PMID:23029612

  1. Utility of PCR in diagnosing pulmonary tuberculosis.

    PubMed Central

    Bennedsen, J; Thomsen, V O; Pfyffer, G E; Funke, G; Feldmann, K; Beneke, A; Jenkins, P A; Hegginbothom, M; Fahr, A; Hengstler, M; Cleator, G; Klapper, P; Wilkins, E G

    1996-01-01

    At present, the rapid diagnosis of pulmonary tuberculosis rests with microscopy. However, this technique is insensitive and many cases of pulmonary tuberculosis cannot be initially confirmed. Nucleic acid amplification techniques are extremely sensitive, but when they are applied to tuberculosis diagnosis, they have given variable results. Investigators at six centers in Europe compared a standardized PCR system (Amplicor; Roche) against conventional culture methods. Defined clinical information was collected. Discrepant samples were retested, and inhibition assays and backup amplification with a separate primer pair were performed. Mycobacterium tuberculosis complex organisms were recovered from 654 (9.1%) of 7,194 samples and 293 (7.8%) of 3,738 patients. Four hundred fifty-two of the M. tuberculosis isolates from 204 patients were smear positive and culture positive. Among the culture-positive specimens, PCR had a sensitivity of 91.4% for smear-positive specimens and 60.9% for smear-negative specimens, with a specificity of 96.1%. Analysis of 254 PCR-positive, culture-negative specimens with discrepant results revealed that 130 were from patients with recently diagnosed tuberculosis and 94 represented a presumed laboratory error. Similar analysis of 118 PCR-negative, culture-positive specimens demonstrated that 27 discrepancies were due to presumed uneven aliquot distribution and 11 were due to presumed laboratory error; PCR inhibitors were detected in 8 specimens. Amplicor enables laboratories with little previous experience with nucleic acid amplification to perform PCR. Disease in more than 60% of the patients with tuberculosis with smear-negative, culture-positive specimens can be diagnosed at the time of admission, and potentially all patients with smear-positive specimens can immediately be confirmed as being infected with M. tuberculosis, leading to improved clinical management. PMID:8735089

  2. Effect of Antiviral Therapy on Serum Activity of Angiotensin Converting Enzyme in Patients with Chronic Hepatitis C

    PubMed Central

    Husic-Selimovic, Azra; Sofic, Amela; Huskic, Jasminko; Bulja, Deniz

    2016-01-01

    Introduction: Renin-angiotenzin system (RAS) is frequently activated in patients with chronic liver disease. Angiotenzin - II (AT-II), produced by angiotenzin converting enzyme (ACE), has many physiological effects, including an important role in liver fibrogenesis. Combined antiviral therapy with PEG-IFN and ribavirin besides its antiviral effect also leads to a reduction in liver parenchyma fibrosis. Aim of the study: Determining the value of ACE in serum of patients with chronic hepatitis C before and after combined antiviral therapy, as well as the value of ACE activities in sera of the control group. Materials and methods: We studied 50 patients treated at Gastroenterohepatology Department, in the time-period of four years. Value of ACE in serum was determined by Olympus AU 400 device, with application of kit “Infinity TN ACE Liquid Stable Reagent”. HCV RNA levels in sera were measured by real time PCR. HCV RNA test was performed with modular analysis of AMPLICOR and COBAS AMPLICOR HCV MONITOR test v2.0, which has proved infection and was used for quantification of the viruses and monitoring of the patients’ response to therapy. Liver histology was evaluated in accordance with the level of necroinflammation activity and stage of fibrosis. Results: Serum activities of ACE in chronic hepatitis C patients is statistically higher than the values in the control group (p=0.02). Antiviral therapy in chronic hepatitis C patients statistically decreases serum activities of ACE (p= 0.02) and indirectly affects fibrogenesis of the liver parenchyma. Correlation between ACE and ALT activity after the therapy was proved (0.3934). Conclusion: Our findings suggest that the activity of ACE in serum is a good indirect parameter of the liver damage, and could be used as an indirect prognostic factor of the level of liver parenchyma damage. Serum activity of ACE can be used as a parameter for non-invasive assessment of intensity of liver damage. PMID:27147779

  3. Development and verification of an automated sample processing protocol for quantitation of human immunodeficiency virus type 1 RNA in plasma.

    PubMed

    Lee, Brenda G; Fiebelkorn, Kristin R; Caliendo, Angela M; Nolte, Frederick S

    2003-05-01

    We developed and verified an automated sample processing protocol for use with the AMPLICOR HIV-1 MONITOR test, version 1.5 (Roche Diagnostics, Indianapolis, Ind.). The automated method uses the MagNA Pure LC instrument and total nucleic acid reagents (Roche Applied Science, Indianapolis, Ind.) to extract human immunodeficiency virus type 1 (HIV-1) RNA from plasma specimens. We compared the HIV-1 load results for a dilution series (1 to 5 nominal log(10) copies/ml) and 175 clinical specimens processed by the automated method to those for the same samples processed by the manual methods specified by the manufacturer. The sensitivity, dynamic range, and precision of the viral load assay obtained by automated processing of specimens were similar to those obtained by an ultrasensitive manual processing method. The results were highly correlated (R(2), 0.95), and were in close agreement, with a mean difference of 0.09 log(10) (standard deviation, 0.292). The limits of agreement were +/-0.58 log(10) for results for samples processed by both the manual and the automated methods. These performance characteristics were achieved with a smaller sample volume (200 versus 500 microl) and without a high-speed centrifugation step and required only 15 min of labor for a batch of 32 samples. In conclusion, the automated sample preparation protocol can replace both the standard and the ultrasensitive manual methods used with the AMPLICOR HIV-1 MONITOR test and can substantially reduce the labor associated with this test. PMID:12734249

  4. Hepatitis C virus RNA assays: a comparison of SuperQuant and Monitor.

    PubMed

    Hadziyannis, E; Hadziyannis, A; Yen-Lieberman, B; Kiwi, M L; Hodnick, S; Spanou, F; Starkey, C; Younossi, Z M

    2001-07-01

    Hepatitis C RNA testing has been used extensively to assess the efficacy of antiviral therapy and has increasingly become an integral part of clinical management of patients with chronic hepatitis C. A variety of commercially available hepatitis C virus (HCV) RNA tests are used to detect HCV RNA qualitatively or quantitatively. These commercial tests have fundamental differences that are reflected on the values they generate. We compared two widely used assays, HCV SuperQuant (SQ) and Amplicor HCV Monitor (M1 and M2), in sera of patients with chronic hepatitis C. A total of 506 sera from 79 patients were tested with both assays. The data were logarithmically transformed and analyzed by linear regression and measurement of agreement. Two hundred thirty-eight sera had HCV RNA values within the dynamic range of both assays. The correlation between the assays was fair, with a correlation coefficient (r) of 0.699. Overall, SQ generated higher values than M1 with a mean difference of 0.558 log (SD = 0.624). One hundred ninety-four (38%) and 121 (24%) of the sera were below the dynamic range of M1 and SQ, respectively. Seventy-three sera, undetectable by M1, were positive by SQ. The Amplicor HCV Monitor 2.0 (M2) was performed in 66 sera. All were positive by SQ and M2, but only 38 were within the dynamic range of M1. The correlations between these tests were good (r = 0.68-0.78), but the agreement was rather poor. In conclusion, this study confirms that both SQ and M2 are more sensitive than M1. Additionally, our results show rather poor agreements between these assays. The recent attempts in standardizing the reporting of these assays should make their results more easily interchangeable. PMID:11418790

  5. Evaluating HPV-negative CIN2+ in the ATHENA trial.

    PubMed

    Petry, Karl Ulrich; Cox, J Thomas; Johnson, Kristin; Quint, Wim; Ridder, Ruediger; Sideri, Mario; Wright, Thomas C; Behrens, Catherine M

    2016-06-15

    A post hoc analysis of the ATHENA study was performed to determine whether true HPV-negative cervical lesions occur and whether they have clinical relevance. The ATHENA database was searched for all CIN2 or worse (CIN2+) cases with cobas HPV-negative results and comparison was made with Linear Array (LA) and Amplicor to detect true false-negative HPV results. Immunostaining with p16 was performed on these cases to identify false-positive histology results. H&E slides were re-reviewed by the study pathologists with knowledge of patient age, HPV test results and p16 immunostaining. Those with positive p16 immunostaining and/or a positive histopathology review underwent whole tissue section HPV PCR by the SPF10/LiPA/RHA system. Among 46,887 eligible women, 497 cases of CIN2+ were detected, 55 of which tested negative by the cobas(®) HPV Test (32 CIN2, 23 CIN3/ACIS). By LA and/or Amplicor, 32 CIN2+ (20 CIN2, 12 CIN3/ACIS) were HPV positive and categorized as false-negatives by cobas HPV; nine of 12 false-negative CIN3/ACIS cases were p16+. There were 23 cases (12 CIN2, 11 CIN3/ACIS) negative by all HPV tests; seven of 11 CIN3/ACIS cases were p16+. H&E slides were available for six cases for re-review and all were confirmed as CIN3/ACIS. Tissue PCR was performed on the six confirmed CIN3/ACIS cases (and one without confirmation): four were positive for HPV types not considered oncogenic, two were positive for oncogenic genotypes and one was indeterminate. In summary, subanalysis of a large cervical cancer screening study did not identify any true CIN3/ACIS not attributable to HPV. PMID:26851121

  6. Evaluation of New Quantitative Assays for Diagnosis and Monitoring of Cytomegalovirus Disease in Human Immunodeficiency Virus-Positive Patients

    PubMed Central

    Pellegrin, Isabelle; Garrigue, Isabelle; Binquet, Christine; Chene, Genevieve; Neau, Didier; Bonot, Pascal; Bonnet, Fabrice; Fleury, Herve; Pellegrin, Jean-Luc

    1999-01-01

    Cobas Amplicor CMV Monitor (CMM) and Quantiplex CMV bDNA 2.0 (CMV bDNA 2.0), two new standardized and quantitative assays for the detection of cytomegalovirus (CMV) DNA in plasma and peripheral blood leukocytes (PBLs), respectively, were compared to the CMV viremia assay, pp65 antigenemia assay, and the Amplicor CMV test (P-AMP). The CMV loads were measured in 384 samples from 58 human immunodeficiency virus (HIV) type 1-infected, CMV-seropositive subjects, including 13 with symptomatic CMV disease. The assays were highly concordant (agreement, 0.88 to 0.97) except when the CMV load was low. Quantitative results for plasma and PBLs were significantly correlated (Spearman ρ = 0.92). For PBLs, positive results were obtained 125 days before symptomatic CMV disease by CMV bDNA 2.0 and 124 days by pp65 antigenemia assay, whereas they were obtained 46 days before symptomatic CMV disease by CMM and P-AMP. At the time of CMV disease diagnosis, the sensitivity, specificity, and positive and negative predictive values of CMV bDNA 2.0 were 92.3, 97.8, 92.3, and 97.8%, respectively, whereas they were 92.3, 93.3, 80, and 97.8%, respectively, for the pp65 antigenemia assay; 84.6, 100, 100, and 95.7%, respectively, for CMM; and 76.9, 100, 100, and 93.8%, respectively, for P-AMP. Considering the entire follow-up, the sensitivity, specificity, and positive and negative predictive values of CMV bDNA 2.0 were 92.3, 73.3, 52.1, and 97.1%, respectively, whereas they were 100, 55.5, 39.4, and 100%, respectively, for the pp65 antigenemia assay; 92.3, 86.7, 66.7, and 97.5%, respectively, for CMM; and 84.6, 91.1, 73.3, and 95.3%, respectively, for P-AMP. Detection of CMV in plasma is technically easy and, despite its later positivity (i.e., later than in PBLs), can provide enough information sufficiently early so that HIV-infected patients can be effectively treated. In addition, these standardized quantitative assays accurately monitor the efficacy of anti-CMV treatment. PMID:10488165

  7. Single-step PCR in molecular diagnosis of hepatitis C virus infection.

    PubMed Central

    Farma, E; Boeri, E; Bettini, P; Repetto, C M; McDermott, J; Lillo, F B; Varnier, O E

    1996-01-01

    The diagnostic utility of two PCR systems and three PCR detection methods for hepatitis C virus (HCV) RNA was evaluated in serum samples. A nested PCR was considered the reference assay and was compared with two single-step PCR methods: the first is based on the detection of PCR products by liquid hybridization with a 32P-end-labeled probe, and the second is the Roche Amplicor colorimetric assay using microwell plate hybridization with a specific nucleic acid probe. Using the Pelicheck HCV RNA Eurohep genotype 1 proficiency panel, our laboratory achieved medium-high levels of performance with all three methods. The highest sensitivity was, however, observed with the isotopic single-step PCR (ss-PCR) method. The analytical sensitivity of ss-PCR with isotopic detection and ss-PCR with colorimetric detection was identical to that of nested PCR, with a 100% result concordance. Comparison of ss-PCR with enzyme-linked immunosorbent and RIBA assays in the analysis of clinical samples showed a high concordance. ss-PCR methods appear more suitable for diagnostic application. Nevertheless, HCV RNA PCR cannot be considered a screening assay; it should be requested in the presence of reactive serology or specific clinical symptomatology with altered liver parameters, and it is a potential tool for the follow-up of patients with HCV infection. PMID:8940466

  8. Evaluation of in-house polymerase chain reaction assay sensitivity, can it be utilized in limited-resources settings?

    PubMed Central

    Dorudinia, Atosa; Shamaei, Masoud; Karimi, Shirin; Javadi, Alireza; Mohammadi Ziazi, Leila; Pourabdollah, Mihan

    2014-01-01

    Background: Polymerase chain reaction (PCR) assay has widely used for the detection of tuberculosis (TB). This study tried to compare in-house PCR with some well-known commercial PCR kits for detection of TB agent. Methods: Clinical samples obtained from 620 TB suspected patients were analyzed for the diagnosis of Mycobacterium tuberculosis complex (MTC) by in-house PCR. All samples were obtained through pulmonary specimens consisted of 384 sputum, 148 bronchial aspirates and 88 pleural effusions. Results: Considering culture as a golden criterion, in which its diagnostic sensitivity and specificity of PCR assay were 87.7% and 85.6%, respectively. The findings of this study also indicate 22.1% (137/620) of the specimens were detected as MTC by PCR. Both PCR and culture confirmed presence of MTC in 57 of the samples. In comparison to culture, the diagnostic sensitivity of PCR for sputum was 87.5% (42/48), bronchial aspirates 100% (12/12), and 60% (3/5) for pleural effusions. The sensitivity of in-house PCR method is comparable with the sensitivity of Amplicor and Cobas TaqMan for MTC. Conclusion: The study illustrates the in-house PCR assay for detection of MTC has high sensitivity and specificity versus approved commercial kits. This could be reliable test in the diagnosis of MTC in resource-limited countries. PMID:25679005

  9. Urogenital Mycoplasmas and Human Papilloma Virus in Hemodialysed Women

    PubMed Central

    Ekiel, Alicja; Pietrzak, Bronisława; Aptekorz, Małgorzata; Mazanowska, Natalia; Kamiński, Paweł; Martirosian, Gayane

    2013-01-01

    Bacterial infections, especially endogenous, are the frequent complications among hemodialyzed and renal transplant patients. In this study we assumed the prevalence of urogenital mycoplasmas and HPV among hemodialysed women. We examined 32 hemodialysed women aged 20–48 (mean 35.6 ± 8.23) and 100 healthy controls of the same ages. Two swabs were collected for detection of mycoplasmas and HPV. Culture of Ureaplasma spp. and M. hominis was performed using Mycoplasma IST2 (bioMérieux, France), Identificaton of U. parvum and U. urealyticum was performed by Kong. Primers described by Jensen were used for M. genitalium. For detection of high-risk HPV types Amplicor HPV (Roche Molecular System, CA) was used. Prevalence of urogenital mycoplasmas in the hemodialysed women (53.1%) was significantly higher (P = 0.0059), compared with controls (25%). In both groups, U. parvum was the most frequently isolated. Cooccurrence of urogenital mycoplasmas was shown in 75% of the HPV-positive hemodialysed women and in 30.4% of HPV-positive controls (P = 0.0461). Cooccurrence of urogenital mycoplasmas with HPV was significantly higher in hemodialysed women. The need to take into account these microorganisms in routine diagnostic, especially for hemodialysed patients, was demonstrated. Further studies to demonstrate the role of this cooccurrence in etiopathogenesis of infection in hemodialysed patients are required. PMID:24363622

  10. Single genome amplification of proviral HIV-1 DNA from dried blood spot specimens collected during early infant screening programs in Lusaka, Zambia

    PubMed Central

    Seu, Lillian; Mwape, Innocent; Guffey, M. Bradford

    2014-01-01

    The ability to evaluate individual HIV-1 virions from the quasispecies of vertically infected infants was evaluated in a field setting at the Centre for Infectious Disease Research in Zambia. Infant heel-prick blood specimens were spotted onto dried blood spot (DBS) filter paper cards at government health clinics. Nucleic acid was extracted and used as a template for HIV-1 proviral DNA detection by a commercial Amplicor HIV-1 PCR test (Roche, version 1.5). On samples that tested positive by commercial diagnostic assay, amplification of DNA was performed using an in-house assay of the 5′ and 3′ region of the HIV-1 genome. Additionally, fragments covering 1200 nucleotides within pol (full length protease and partial reverse transcriptase) and 1400 nucleotides within env (variable 1-variable 5 region) were further analyzed by single genome amplification (SGA). In summary, we have demonstrated an in-house assay for amplifying the 5′ and 3′ proviral HIV-1 DNA as well as pol and env proviral DNA fragments from DBS cards collected and analyzed entirely in Zambia. In conclusion, this study shows the feasibility of utilizing DBS cards to amplify the whole proviral HIV-1 genome as well as perform SGA on key HIV-1 genes. PMID:24667303

  11. Urogenital mycoplasmas and human papilloma virus in hemodialysed women.

    PubMed

    Ekiel, Alicja; Pietrzak, Bronisława; Wiechuła, Barbara; Aptekorz, Małgorzata; Mazanowska, Natalia; Rady, Dominika; Kamiński, Paweł; Martirosian, Gayane

    2013-01-01

    Bacterial infections, especially endogenous, are the frequent complications among hemodialyzed and renal transplant patients. In this study we assumed the prevalence of urogenital mycoplasmas and HPV among hemodialysed women. We examined 32 hemodialysed women aged 20-48 (mean 35.6 ± 8.23) and 100 healthy controls of the same ages. Two swabs were collected for detection of mycoplasmas and HPV. Culture of Ureaplasma spp. and M. hominis was performed using Mycoplasma IST2 (bioMérieux, France), Identificaton of U. parvum and U. urealyticum was performed by Kong. Primers described by Jensen were used for M. genitalium. For detection of high-risk HPV types Amplicor HPV (Roche Molecular System, CA) was used. Prevalence of urogenital mycoplasmas in the hemodialysed women (53.1%) was significantly higher (P = 0.0059), compared with controls (25%). In both groups, U. parvum was the most frequently isolated. Cooccurrence of urogenital mycoplasmas was shown in 75% of the HPV-positive hemodialysed women and in 30.4% of HPV-positive controls (P = 0.0461). Cooccurrence of urogenital mycoplasmas with HPV was significantly higher in hemodialysed women. The need to take into account these microorganisms in routine diagnostic, especially for hemodialysed patients, was demonstrated. Further studies to demonstrate the role of this cooccurrence in etiopathogenesis of infection in hemodialysed patients are required. PMID:24363622

  12. Diagnosis of Neisseria gonorrhoeae Using Molecular Beacon

    PubMed Central

    Patel, Achchhe Lal; Sonkar, Subash Chandra; Kumari, Indu; Saluja, Daman

    2015-01-01

    Neisseria gonorrhoeae is an important sexually transmitted diseases (STD) causing pathogen worldwide. Due to absence of an affordable diagnostic assay, routine screening of gonococcal infection becomes impossible in developing countries where infection rates are maximum. Treatment is given on the basis of symptoms alone which leads to spread of infection. Thus, development of a rapid, sensitive, specific, and PCR based visual diagnostic assay suitable for developing countries, required for better disease management, is aimed at in present study. Endocervical swabs were collected from patients visiting gynecology department of various hospitals in Delhi. In-house PCR based assay was developed and modified to visual assay using molecular beacon for end-point detection. It was evaluated against Roche AMPLICOR NG kit and rmp gene. Specificity of beacon was confirmed by competition experiments. Diagnostic test was 98.21% specific and 99.59% sensitive whereas negative and positive predicted value were 99.40% and 98.78%, respectively. We also observed that twice the concentration (2X) of premix was stable at 4°C for 4 months and dry swab samples gave concordant results with that of wet swabs. These features make the test best suitable for routine diagnosis of genital infections in developing countries. PMID:25802857

  13. Validation of the Gen-Probe Aptima qualitative HIV-1 RNA assay for diagnosis of human immunodeficiency virus infection in infants.

    PubMed

    Fiscus, Susan A; McMillion, Takesha; Nelson, Julie A E; Miller, William C

    2013-12-01

    The qualitative Roche HIV-1 DNA Amplicor assay has been used for the past 20 years to diagnose HIV infection in infants and young children but is being phased out; hence, alternative assays must be found. The Gen-Probe Aptima qualitative HIV-1 RNA assay is currently the only FDA-cleared HIV-1 nucleic acid assay approved for diagnosis, but data on the use of this assay with infant plasma are limited. We assessed Aptima's performance using control material for reproducibility and limit of detection and 394 plasma samples (0.2 to 0.5 ml) from HIV-exposed infected and uninfected infants and children for analytical sensitivity and specificity. Assays to assess within-run repeatability and between-run reproducibility indicated that the controls with 10,000 (5 of 5), 200 (5 of 5), 100 (16 of 16), 50 (12 of 12), and 25 (20 of 20) HIV-1 RNA copies/ml (cp/ml) were always positive, and negatives were always negative (20 of 20). The limit of detection was 14 cp/ml, as determined by probit analysis. The analytic sensitivity of the assay was 99.5% (189/190 samples; 95% confidence interval [CI], 97.1 to 99.9%) and specificity was 99.5% (199/200 samples; 95% CI, 97.2 to 99.9%). These results suggest that the assay is suitable for early infant diagnosis of HIV-1. PMID:24088864

  14. Geospatial Distribution and Clustering of Chlamydia trachomatis in Communities Undergoing Mass Azithromycin Treatment

    PubMed Central

    Yohannan, Jithin; He, Bing; Wang, Jiangxia; Greene, Gregory; Schein, Yvette; Mkocha, Harran; Munoz, Beatriz; Quinn, Thomas C.; Gaydos, Charlotte; West, Sheila K.

    2014-01-01

    Purpose. We detected spatial clustering of households with Chlamydia trachomatis infection (CI) and active trachoma (AT) in villages undergoing mass treatment with azithromycin (MDA) over time. Methods. We obtained global positioning system (GPS) coordinates for all households in four villages in Kongwa District, Tanzania. Every 6 months for a period of 42 months, our team examined all children under 10 for AT, and tested for CI with ocular swabbing and Amplicor. Villages underwent four rounds of annual MDA. We classified households as having ≥1 child with CI (or AT) or having 0 children with CI (or AT). We calculated the difference in the K function between households with and without CI or AT to detect clustering at each time point. Results. Between 918 and 991 households were included over the 42 months of this analysis. At baseline, 306 households (32.59%) had ≥1 child with CI, which declined to 73 households (7.50%) at 42 months. We observed borderline clustering of households with CI at 12 months after one round of MDA and statistically significant clustering with growing cluster sizes between 18 and 24 months after two rounds of MDA. Clusters diminished in size at 30 months after 3 rounds of MDA. Active trachoma did not cluster at any time point. Conclusions. This study demonstrates that CI clusters after multiple rounds of MDA. Clusters of infection may increase in size if the annual antibiotic pressure is removed. The absence of growth after the three rounds suggests the start of control of transmission. PMID:24906862

  15. The Effect of Multiple Rounds of Mass Drug Administration on the Association between Ocular Chlamydia trachomatis Infection and Follicular Trachoma in Preschool-Aged Children

    PubMed Central

    Lee, Jennifer S.; Muñoz, Beatriz E.; Mkocha, Harran; Gaydos, Charlotte A.; Quinn, Thomas C.; West, Sheila K.

    2014-01-01

    Purpose To examine the relationship between ocular Chlamydia trachomatis infection and follicular trachoma (TF) in children prior to and following multiple rounds of annual mass drug administration (MDA) with azithromycin. Methodology/principal findings Thirty-two communities with endemic trachoma in Kongwa District, Tanzania, were offered annual MDA as part of a district-wide trachoma control program. Presence of ocular C. trachomatis infection and TF were assessed in 3,200 randomly sampled children aged five years and younger, who were examined prior to each MDA. Infection was detected using the Amplicor CT/NG assay and TF was identified by clinical examination using the World Health Organization (WHO) simplified grading system. The association between chlamydial infection and TF in children was evaluated at baseline prior to any treatment, and 12 months after each of three annual rounds of mass treatment. Factors associated with infection were examined using generalized estimating equation models. At baseline, the overall prevalence of chlamydial infection and TF was 22% and 31%, respectively. Among children with clinical signs of TF, the proportion of those with infection was 49% prior to treatment and declined to 30% after three MDAs. The odds of infection positivity among children with clinical signs of TF decreased by 26% (OR 0.74, 95% CI 0.65 to 0.84, p = <0.01) with each MDA, after adjusting for age. For children aged under one year, who did not receive treatment, the relationship was unchanged. Conclusions/significance The association between ocular C. trachomatis infection and TF weakened in children with each MDA, as both infection and clinical disease prevalence declined. However, there was still a significant proportion of TF cases with infection after three rounds of MDA. New strategies are needed to assess this residual infection for optimal treatment distribution. PMID:24722392

  16. Ag(I)-coordinated hairpin DNA for homogenous electronic monitoring of hepatitis C virus accompanying isothermal cycling signal amplification strategy.

    PubMed

    Lu, Minghua; Xu, Linfang; Zhang, Xiaona; Xiao, Rui; Wang, Youmei

    2015-11-15

    This work designs a new homogenous electronic monitoring platform for sensitive detection of hepatitis C virus (HCV) on an immobilization-free Ag(I)-assisted hairpin DNA through the cytosine-Ag(+)-cytosine coordination chemistry. The assay consists of target-induced Ag(+) dissociation from hairpin DNA and an isothermal circular strand-displacement polymerization (ICSDP) reaction. Upon target analyte introduction, HCV DNA initially hybridizes with hairpin DNA to disrupt the Ag(I)-coordinated hairpin probe and releases the coordinated Ag(+) ion, then the newly formed DNA duplex induces the ICSDP reaction with the aid of primer and polymerase, and then the displaced target DNA retriggers Ag(I)-coordinated hairpin DNA with target recycling, thereby resulting in formation of numerous free Ag(+) ions in the detection cell. The released Ag(+) ions can be readily captured by the negatively charged screen-printed carbon electrode, and subsequent anodic-stripping voltammetric detection of the captured Ag(+) ions are conducted to form the anodic current for the production of the electrochemical signal within the applied potential. Under optimal conditions, the ICSDP-based homogenous sensing system can be utilized for the detection of HCV DNA at a concentration as low as 2.3 pM. Intra- and inter-assay coefficients of variation with identical batches are below 9.5% and 10.5%, respectively. The analysis in 5 clinical serum specimens shows good accordance between results obtained by the developed method and commercial Cobas® Amplicor HCV Test Analyzer. PMID:26071691

  17. Elevated Cytokine and Chemokine Levels in the Placenta Are Associated With in utero HIV-1 Mother-To-Child Transmission

    PubMed Central

    Kumar, Surender B.; Rice, Cara E.; Milner, Danny A.; Ramirez, Nilsa C.; Ackerman, William E.; Mwapasa, Victor; Turner, Abigail Norris; Kwiek, Jesse J.

    2012-01-01

    Objective To determine if there is an association between cytokine and chemokine levels in plasma isolated from the placenta and HIV-1 mother-to-child transmission (MTCT). Design We designed a case-control study of HIV-infected, pregnant women enrolled in the Malaria and HIV in Pregnancy cohort. Participants were recruited in Blantyre, Malawi from 2000-04. Cases were women whose children were HIV-1 DNA-positive at birth (in utero MTCT) or HIV-1 DNA-negative at birth and HIV-1 DNA-positive at 6-weeks post-partum (intrapartum MTCT); controls were women whose children were HIV-1 DNA-negative both at birth and 6-weeks post-partum. Methods After delivery, blood was isolated from an incision on the basal plate of the placenta. We used a Luminex assay to simultaneously quantify 27 cytokines, chemokines, and growth factors in placental plasma. HIV-1 RNA copies were quantified with the Roche Amplicor kit. Results Levels of IL-4, IL-5, IL-6, IL-7, IL-9, eotaxin, IL1Ra and IP-10 were significantly elevated in placental plasma isolated from cases of in utero HIV-1 MTCT. In contrast, only GCSF was elevated in placental plasma isolated from cases of intrapartum MTCT. After adjusting for maternal age, gestational age, and peripheral CD4+ T cell count, every log10 increase in placental IP-10 was associated with a three-fold increase in the prevalence of in utero HIV-1 MTCT. Conclusions Elevated cytokine and chemokine levels in placental plasma were associated with in utero and not intrapartum MTCT. IP-10, which is both a T-cell chemokine and potentiator of HIV-replication, was robustly and independently associated with prevalent, in utero MTCT. PMID:22301415

  18. Rapid Quantification of Hepatitis B Virus DNA by Automated Sample Preparation and Real-Time PCR

    PubMed Central

    Stelzl, Evelyn; Muller, Zsofia; Marth, Egon; Kessler, Harald H.

    2004-01-01

    Monitoring of hepatitis B virus (HBV) DNA in serum by molecular methods has become the standard for assessment of the replicative activity of HBV. Several molecular assays for the detection and quantification of HBV DNA have been described. However, they usually lack automated sample preparation. Moreover, those assays, which are based on PCR, are limited by a short dynamic range (2 to 3 log units). In the present study, the use of RealArt HBV LC PCR Reagents in conjunction with automated extraction on the COBAS AMPLIPREP analyzer was evaluated. Members of an HBV proficiency program panel were tested; linearity, interassay, and intra-assay variations were determined. The performance of the assay in a routine clinical laboratory was evaluated with a total of 117 clinical specimens. When members of the HBV proficiency program panel were tested by the new molecular assay, the results were found to be within ±0.5 log unit of the results obtained by reference laboratories. Determination of linearity resulted in a quasilinear curve over more than 6 log units. The interassay variation of the RealArt HBV LC PCR Reagents by use of the automated sample preparation protocol ranged from 16 to 73%, and the intra-assay variation ranged from 9 to 40%. When clinical samples were tested by the new assay with the automated sample preparation protocol and the results were compared with those obtained by the COBAS AMPLICOR HBV MONITOR Test with manual sample preparation, the results for 76% of all samples with positive results by both tests were found to be within ±0.5 log unit and the results for another 18% were found to be within between 0.5 and 1.0 log unit. In conclusion, the real-time PCR assay with automated sample preparation proved to be suitable for the routine molecular laboratory and required less hands-on time. PMID:15184417

  19. Abbott RealTime PCR assay is useful for evaluating virological response to antiviral treatment for chronic hepatitis C.

    PubMed

    Ikezaki, Hiroaki; Furusyo, Norihiro; Ihara, Takeshi; Hayashi, Takeo; Ogawa, Eiichi; Toyoda, Kazuhiro; Taniai, Hiroaki; Kainuma, Mosaburo; Murata, Masayuki; Hayashi, Jun

    2011-12-01

    This study was done to evaluate the utility of the Abbott RealTime PCR assay (ART) for the monitoring of chronic hepatitis C patients. The serum samples of 183 patients infected with hepatitis C virus (HCV) genotype 1b who had completed a 48-week period of pegylated interferon (PEG-IFN) alpha-2b plus ribavirin treatment were prospectively analyzed. Serum HCV RNA levels were measured both by ART and by the Roche COBAS Amplicor Monitor test, version2.0 (CAM) at baseline and at weeks 4, 12, 24, 36, and 48 of treatment, and at 24 weeks after the end of treatment (EOT). A significant positive correlation of pretreatment HCV RNA levels was found between ART and CAM (r = 0.595, P < 0.0001). Of the 183 patients, 66 (36.0%) achieved a sustained virological response (SVR). The logarithmic decline of the HCV RNA level from the pretreatment level determined by ART in SVR patients was significantly higher than that in non-SVR patients at all time points tested. The logarithmic decline determined by CAM in SVR patients was significantly higher than that in non-SVR patients only at week 4, but there was no significant difference at other weeks. Of 124 patients who were HCV RNA-negative at EOT by ART, 58 (46.8%) had a relapse of viremia at 24 weeks after EOT, whereas 77 of 143 patients (53.8%) who were HCV RNA-negative at EOT by CAM had a relapse. The relapse rate was lower when determined by ART than by CAM, but not significantly so. ART is more useful than CAM for evaluating the virological response to antiviral treatment for chronic hepatitis C. PMID:21528383

  20. Magnitude of Virologic Blips Is Associated With a Higher Risk for Virologic Rebound in HIV-Infected Individuals: A Recurrent Events Analysis

    PubMed Central

    Grennan, J. Troy; Loutfy, Mona R.; Su, DeSheng; Harrigan, P. Richard; Cooper, Curtis; Klein, Marina; Machouf, Nima; Montaner, Julio S. G.; Rourke, Sean; Tsoukas, Christos; Hogg, Bob

    2012-01-01

    (See the editorial commentary by Taiwo and Bosch, on pages 1189–91.) Background. The importance of human immunodeficiency virus (HIV) blip magnitude on virologic rebound has been raised in clinical guidelines relating to viral load assays. Methods. Antiretroviral-naive individuals initiating combination antiretroviral therapy (cART) after 1 January 2000 and achieving virologic suppression were studied. Negative binomial models were used to identify blip correlates. Recurrent event models were used to determine the association between blips and rebound by incorporating multiple periods of virologic suppression per individual. Results. 3550 participants (82% male; median age, 40 years) were included. In a multivariable negative binomial regression model, the Amplicor assay was associated with a lower blip rate than branched DNA (rate ratio, 0.69; P < .01), controlling for age, sex, region, baseline HIV-1 RNA and CD4 count, AIDS-defining illnesses, year of cART initiation, cART type, and HIV-1 RNA testing frequency. In a multivariable recurrent event model controlling for age, sex, intravenous drug use, cART start year, cART type, assay type, and HIV-1 RNA testing frequency, blips of 500–999 copies/mL were associated with virologic rebound (hazard ratio, 2.70; P = .002), whereas blips of 50–499 were not. Conclusions. HIV-1 RNA assay was an important determinant of blip rates and should be considered in clinical guidelines. Blips ≥500 copies/mL were associated with increased rebound risk. PMID:22438396

  1. Dried-Plasma Transport Using a Novel Matrix and Collection System for Human Immunodeficiency Virus and Hepatitis C Virus Virologic Testing▿

    PubMed Central

    Lloyd, R. M.; Burns, D. A.; Huong, J. T.; Mathis, R. L.; Winters, M. A.; Tanner, M.; De La Rosa, A.; Yen-Lieberman, B.; Armstrong, W.; Taege, A.; McClernon, D. R.; Wetshtein, J. L.; Friedrich, Brian M.; Ferguson, Monique R.; O'Brien, William; Feorino, P. M.; Holodniy, M.

    2009-01-01

    A novel method for the collection and transportation of dried-blood-plasma samples, SampleTanker (ST), was developed and compared to standard shipping protocols for frozen-plasma specimens containing human immunodeficiency virus type 1 (HIV-1) and/or hepatitis C virus (HCV). Matched frozen and dried 1-ml EDTA-containing plasma samples were collected and analyzed by several molecular-based virologic assays. After addition of 1.175 ml of reconstitution buffer, 1.035 ml of dried plasma was recovered. Mean intra-assay variances were 0.05, 0.05, and 0.06 log10 copies/ml for the Versant, Amplicor, and NucliSens QT HIV-1 load assays, respectively (P, not significant). However, mean HIV-1 viral load was consistently reduced in dried samples by 0.32 to 0.51 log10 copies/ml, depending on assay type (P < 0.05). Infectious HIV-1 was not recovered from dried ST plasma. There was no significant difference in HIV-1 viral load results obtained using ST after 8 weeks of storage at ambient temperature. Compared to frozen plasma, HIV-1 genotypic results were >99% concordant at the nucleotide and amino acid levels, as well as for resistance-associated mutations. We further demonstrated successful detection of multiple analytes, including HIV-1 viral load, HIV-1 antiretroviral resistance genotype, and HCV genotype, from a single ST unit. Dried plasma collected with ST yielded comparable results to frozen samples for multiple-analyte clinical testing. As such, ST could be a useful alternative for virologic tests and clinical trials worldwide by significantly diminishing transportation cost and the sample volume restrictions associated with dried-blood-spot technology. PMID:19321732

  2. Detection of endocervical anti-Chlamydia trachomatis immunoglobulin A in pregnant women by a rapid, 6-minute enzyme-linked immunosorbent assay: comparison with PCR and chlamydial antigen detection methods.

    PubMed Central

    Witkin, S S; Bongiovanni, A M; Inglis, S R

    1997-01-01

    There is a need for a rapid, uncomplicated, and inexpensive test for Chlamydia trachomatis infection in women. We evaluated the ability of a 6-min enzyme-linked immunosorbent assay (ELISA) that requires no laboratory equipment (IgA Rapid SeroTest; Savyon Diagnostics) to detect C. trachomatis immunoglobulin A (IgA) in the endocervices of 167 inner-city pregnant women and compared the results with DNA amplification (Amplicor PCR; Roche Diagnostics) and antigen detection (Chlamydiazyme; Abbott Laboratories) performed on the same women. Anti-C. trachomatis IgA was detected in the cervices of 32 women (19.2%). Samples from 23 women (13.8%) were PCR positive, while chlamydial antigen was present in 20 women (12.0%). There was only 1 sample (4.3%) that was positive by PCR but negative by ELISA; 10 samples were ELISA positive and PCR negative. In contrast, seven samples (30.4%) were PCR positive but Chlamydiazyme negative and four were Chlamydiazyme positive and PCR negative. Compared to PCR, the IgA ELISA had a sensitivity of 95.7%, a specificity of 93.1%, a positive predictive value of 68.8%, and a negative predictive value of 99.3%. The antigen assay had a sensitivity of only 69.6%, a specificity of 97.2%, a positive predictive value of 80.0%, and a negative predictive value of 95.2%. In high-risk groups where laboratory testing is not available, or where the patient might not return to obtain her testing result and be treated, the Rapid IgA SeroTest is a viable alternative for detection of cervical C. trachomatis in pregnant women. PMID:9196193

  3. Development and implementation challenges of a quality assured HIV infant diagnosis program in Nigeria using dried blood spots and DNA polymerase chain reaction.

    PubMed

    Audu, Rosemary; Onwuamah, Chika; Salu, Olumuyiwa; Okwuraiwe, Azuka; Ou, Chin-Yih; Bolu, Omotayo; Bond, Kyle B; Diallo, Karidia; Lu, Lydia; Jelpe, Tapdiyel; Okoye, McPaul; Ngige, Evelyn; Vertefeuille, John

    2015-04-01

    Nigeria has one of the highest HIV burdens as well as mother-to-infant transmission rates in the world. A pilot program using polymerase chain reaction (PCR)-based testing of dried blood spot (DBS) specimens was implemented to enable early identification of HIV-infected infants and timely referral and linkage to care. From February 2007 to October 2008, whole blood was collected by finger prick to prepare DBS from infants <18 months presenting in six public mother-and-child health facilities in Lagos, Nigeria. The DBS were tested using the Roche Amplicor HIV-1 DNA Test, v1.5. To monitor laboratory testing quality, all of the PCR-positive and 10% of the PCR-negative DBS were retested by the same method at another reference laboratory. Three hundred and sixty-five randomly selected infants were screened using HIV rapid tests (RT) according to the national algorithm and RT-negative and PCR-positive specimens were also tested using Genscreen enzyme-linked immunosorbent assay (EIA) (Bio-Rad, France). The turnaround time (TAT) from sample collection, testing, and dispatching of results from each health facility was monitored. A total of 1,273 infants with a median age of 12.6 weeks (1 day to 71.6 weeks) participated in the program and 280 (22.0%) were PCR positive. HIV transmission levels varied greatly in the different health facilities ranging from 7.1% to 38.4%. Infants aged 48 to 72 weeks had the highest level of PCR positivity (41.1%). All PCR-positive specimens were confirmed by retesting. The mean turnaround time from DBS collection to returning of the laboratory result to the health facilities was 25 days. Three infants were found to be HIV antibody negative by rapid tests but were positive by both PCR and the fourth generation EIA. The DBS-based PCR program accurately identified all of the HIV-infected infants. However, many programmatic challenges related to the laboratory and TAT were identified. PMID:25381805

  4. Comparison of Three Different FDA-Approved Plasma HIV-1 RNA Assay Platforms Confirms the Virologic Failure Endpoint of 200 Copies per Milliliter Despite Improved Assay Sensitivity

    PubMed Central

    Jennings, Cheryl; Johnson, Victoria A.; Coombs, Robert W.; McKinnon, John E.; Bremer, James W.; Cobb, Bryan R.; Cloherty, Gavin A.; Mellors, John W.; Ribaudo, Heather J.

    2015-01-01

    Discrepancies between HIV-1 RNA results assayed by different FDA-approved platforms have been reported. Plasma samples collected from 332 randomly selected clinical trial participants during the second year of antiretroviral treatment were assayed with three FDA-approved platforms: UltraSensitive Roche Amplicor Monitor, v1.5 (Monitor), the Abbott RealTime HIV-1 test on the m2000 system (Abbott), and the Roche TaqMan HIV-1 test, v2.0 (TaqMan). Samples from 61 additional participants with confirmed HIV-1 RNA levels of >50 copies/ml during trial follow-up were also included. Endpoints were HIV-1 RNA quantification of ≤50 copies/ml versus >50 copies/ml at an individual-sample level (primary) and determination of confirmed virologic failure (VF) from longitudinal samples. A total of 389 participants had results obtained from all assays on at least one sample (median = 6). Proportions of results of >50 copies/ml were 19% (Monitor), 22% (TaqMan), and 25% (Abbott). Despite indication of strong agreement (Cohen's kappa, 0.76 to 0.82), Abbott was more likely to detect HIV-1 RNA levels of >50 copies/ml than Monitor (matched-pair odds ratio [mOR] = 4.2; modified Obuchowski P < 0.001) and TaqMan (mOR = 2.1; P < 0.001); TaqMan was more likely than Monitor (mOR = 2.6; P < 0.001). Despite strong agreement in classifying VF across assay comparisons (kappa, 0.75 to 0.92), at a 50-copies/ml threshold, differences in the probability of VF classification (in the same direction as primary) were apparent (all McNemar's P < 0.007). At a 200-copies/ml VF threshold, no differences between assays were apparent (all P > 0.13). Despite strong agreement among assays, significant differences were observed with respect to detecting HIV-1 RNA levels of >50 copies/ml and identifying VF at the 50-copies/ml threshold. This has important implications for the definition of VF in clinical trials and clinical practice. PMID:26063861

  5. Comparison of Three Different FDA-Approved Plasma HIV-1 RNA Assay Platforms Confirms the Virologic Failure Endpoint of 200 Copies per Milliliter Despite Improved Assay Sensitivity.

    PubMed

    Lalama, Christina M; Jennings, Cheryl; Johnson, Victoria A; Coombs, Robert W; McKinnon, John E; Bremer, James W; Cobb, Bryan R; Cloherty, Gavin A; Mellors, John W; Ribaudo, Heather J

    2015-08-01

    Discrepancies between HIV-1 RNA results assayed by different FDA-approved platforms have been reported. Plasma samples collected from 332 randomly selected clinical trial participants during the second year of antiretroviral treatment were assayed with three FDA-approved platforms: UltraSensitive Roche Amplicor Monitor, v1.5 (Monitor), the Abbott RealTime HIV-1 test on the m2000 system (Abbott), and the Roche TaqMan HIV-1 test, v2.0 (TaqMan). Samples from 61 additional participants with confirmed HIV-1 RNA levels of >50 copies/ml during trial follow-up were also included. Endpoints were HIV-1 RNA quantification of ≤50 copies/ml versus >50 copies/ml at an individual-sample level (primary) and determination of confirmed virologic failure (VF) from longitudinal samples. A total of 389 participants had results obtained from all assays on at least one sample (median = 6). Proportions of results of >50 copies/ml were 19% (Monitor), 22% (TaqMan), and 25% (Abbott). Despite indication of strong agreement (Cohen's kappa, 0.76 to 0.82), Abbott was more likely to detect HIV-1 RNA levels of >50 copies/ml than Monitor (matched-pair odds ratio [mOR] = 4.2; modified Obuchowski P < 0.001) and TaqMan (mOR = 2.1; P < 0.001); TaqMan was more likely than Monitor (mOR = 2.6; P < 0.001). Despite strong agreement in classifying VF across assay comparisons (kappa, 0.75 to 0.92), at a 50-copies/ml threshold, differences in the probability of VF classification (in the same direction as primary) were apparent (all McNemar's P < 0.007). At a 200-copies/ml VF threshold, no differences between assays were apparent (all P > 0.13). Despite strong agreement among assays, significant differences were observed with respect to detecting HIV-1 RNA levels of >50 copies/ml and identifying VF at the 50-copies/ml threshold. This has important implications for the definition of VF in clinical trials and clinical practice. PMID:26063861

  6. Control of Trachoma from Achham District, Nepal: A Cross-Sectional Study from the Nepal National Trachoma Program

    PubMed Central

    Pant, Bidya Prasad; Bhatta, Ramesh C.; Chaudhary, J. S. P.; Awasthi, Suresh; Mishra, Sailesh; Sharma, Shekhar; Cuddapah, Puja A.; Gwyn, Sarah E.; Stoller, Nicole E.; Martin, Diana L.; Keenan, Jeremy D.; Lietman, Thomas M.; Gaynor, Bruce D.

    2016-01-01

    Background The WHO seeks to control trachoma as a public health problem in endemic areas. Achham District in western Nepal was found to have TF (trachoma follicular) above 20% in a 2006 government survey, triggering 3 annual mass drug administrations finishing in 2010. Here we assess the level of control that has been achieved using surveillance for clinical disease, ocular chlamydia trachomatis infection, and serology for antibodies against chlamydia trachomatis protein antigens. Methods We conducted a cross-sectional survey of children aged 1–9 years in communities in Achham District in early 2014 including clinical examination validated with photographs, conjunctival samples for Chlamydia trachomatis (Amplicor PCR), and serological testing for antibodies against chlamydia trachomatis protein antigens pgp3 and CT694 using the Luminex platform. Findings In 24 randomly selected communities, the prevalence of trachoma (TF and/or TI) in 1–9 year olds was 3/1124 (0.3%, 95% CI 0.1 to 0.8%), and the prevalence of ocular chlamydia trachomatis infection was 0/1124 (0%, 95% CI 0 to 0.3%). In 18 communities selected because they had the highest prevalence of trachoma in a previous survey, the prevalence of TF and/or TI was 7/716 (1.0%, 95% CI 0.4 to 2.0%) and the prevalence of ocular chlamydia trachomatis infection was 0/716 (0%, 95% CI 0 to 0.5%). In 3 communities selected for serological testing, the prevalence of trachoma was 0/68 (0%, 95% CI 0 to 5.3%), the prevalence of ocular chlamydia trachomatis infection was 0/68 (0%, 95% CI 0 to 0.5%), the prevalence of antibodies against chlamydia trachomatis protein antigen pgp3 was 1/68 (1.5%, 95% CI 0.04% to 7.9%), and the prevalence of antibodies against chlamydia trachomatis protein antigen CT694 was 0/68 (0%, 95% CI 0 to 5.3%). Conclusion/Significance This previously highly endemic district in Nepal has little evidence of recent clinical disease, chlamydia trachomatis infection, or serological evidence of trachoma

  7. Diagnostic Accuracy of a Prototype Point-of-Care Test for Ocular Chlamydia trachomatis under Field Conditions in The Gambia and Senegal

    PubMed Central

    Harding-Esch, Emma M.; Holland, Martin J.; Schémann, Jean-François; Molina, Sandra; Sarr, Isatou; Andreasen, Aura A.; Roberts, Chrissy h.; Sillah, Ansumana; Sarr, Boubacar; Harding, Edward F.; Edwards, Tansy; Bailey, Robin L.; Mabey, David C. W.

    2011-01-01

    Background The clinical signs of active trachoma are often present in the absence of ocular Chlamydia trachomatis infection in low prevalence and mass treated settings. Treatment decisions are currently based on the prevalence of clinical signs, and this may result in the unnecessary distribution of mass antibiotic treatment. We aimed to evaluate the diagnostic accuracy of a prototype point-of-care (POC) test, developed for field diagnosis of ocular C. trachomatis, in low prevalence settings of The Gambia and Senegal. Methodology/Principal Findings Three studies were conducted, two in The Gambia and one in Senegal. Children under the age of 10 years were screened for the clinical signs of trachoma. Two ocular swabs were taken from the right eye. The first swab was tested by the POC test in the field and the result independently graded by two readers. The second swab was tested for the presence of C. trachomatis by Amplicor Polymerase Chain Reaction. In Senegal, measurements of humidity and temperature in the field were taken. A total of 3734 children were screened, 950 in the first and 1171 in the second Gambian study, and 1613 in Senegal. The sensitivity of the prototype POC test ranged between 33.3–67.9%, the specificity between 92.4–99.0%, the positive predictive value between 4.3–21.0%, and the negative predictive value between 98.0–99.8%. The rate of false-positives increased markedly at temperatures above 31.4°C and relative humidities below 11.4%. Conclusions/Significance In its present format, this prototype POC test is not suitable for field diagnosis of ocular C. trachomatis as its specificity decreases in hot and dry conditions: the environment in which trachoma is predominantly found. In the absence of a suitable test for infection, trachoma diagnosis remains dependent on clinical signs. Under current WHO recommendations, this is likely resulting in the continued mass treatment of non-infected communities. PMID:21829735

  8. Genital human papillomavirus infection in women from the Zagreb region.

    PubMed

    Marijan, Tatjana; Vranes, Jasmina; Mlinarić-Dzepina, Ana; Leskovar, Vladimira; Knezević, Jasna; Kvaternik, Matea

    2007-04-01

    Human papillomavirus (HPV) infection is the most common sexually transmitted infection, especially among young, sexually active individuals. As persistent infection with oncogenic types may lead to cervical cancer, HPV testing is a useful tool to screen for women at risk for subsequent development of cervical cancer. The aim of the study was to determine the prevalence of high-risk HPV (hrHPV) infection in different age groups of cytologically selected women from the Zagreb region, and to evaluate the frequency and results of repeat hrHPV testing. During a one-year study period (November 2005 to November 2006), a total of 3,440 cervical samples from women attending gynecological services of public and private health care systems were received. They were tested for 13 hrHPV genotypes by the polymerase chain reaction based AMPLICOR HPV test (Roche Molecular Systems). The overall prevalence of hrHPV was 34.6%. Most samples were obtained from women aged 21-30 years (44.2%), followed by the 31-40 (27.6%), 41-50 (15.7%), 51-60 (5.3%) and 261 (2.4%) age groups. Out of 3,227 cervical samples obtained from women of known age, 4.9% were obtained from the group of girls younger than 21, in which the highest prevalence of hrHPV (49.4%) was found. A similar prevalence was observed in women aged 21-30 (45.1%). The prevalence gradually decreased with age. During the study period, repeat hrHPV testing was performed in samples from 66 women at different intervals. Out of 28 women that were hrHPV negative on initial testing, only five women turned positive on repeat testing. Out of 38 women that were positive on initial testing, in one-third hrHPV could not be detected on repeat testing. As expected, hrHPV infection was highly prevalent in female adolescents and young women. Further investigation on repeat hrHPV testing is needed to assess virus clearance and rate of newly acquired infection. PMID:17600936

  9. The Diaphragm and Lubricant Gel for Prevention of Cervical Sexually Transmitted Infections: Results of a Randomized Controlled Trial

    PubMed Central

    Ramjee, Gita; van der Straten, Ariane; Chipato, Tsungai; de Bruyn, Guy; Blanchard, Kelly; Shiboski, Stephen; Cheng, Helen; Montgomery, Elizabeth; Padian, Nancy

    2008-01-01

    Background We evaluated the effectiveness of the Ortho All-Flex Diaphragm, lubricant gel (Replens®) and condoms compared to condoms alone on the incidence of chlamydial and gonococcal infections in an open-label randomized controlled trial among women at risk of HIV/STI infections. Methods We randomized 5045 sexually-active women at three sites in Southern Africa. Participants who tested positive for curable STIs were treated prior to enrollment as per local guidelines. Women were followed quarterly and tested for Chlamydia trachomatis (CT) or Neisseria gonorrhoeae (GC) infection by nucleic-acid amplification testing (Roche Amplicor®) using first-catch urine specimens. STIs detected at follow-up visits were treated. We compared the incidence of first infection after randomization between study arms in both intent-to-treat (ITT) and per-protocol populations. Findings Baseline demographic, behavioral and clinical characteristics were balanced across study arms. Nearly 80% of participants were under 35 years of age. Median follow-up time was 21 months and the retention rate was over 93%. There were 471 first chlamydia infections, 247 in the intervention arm and 224 in the control arm with an overall incidence of 6.2/100 woman-years (wy) (relative hazard (RH) 1.11, 95% Confidence Interval (CI): 0.93–1.33; p = 0.25) and 192 first gonococcal infections, 95 in the intervention arm and 97 in the control arm with an overall incidence of 2.4/100wy (RH 0.98, 95%CI: 0.74–1.30; p = 0.90). Per protocol results indicated that when diaphragm adherence was defined as “always use” since the last visit, there was a significant reduction in the incidence of GC infection among women randomized to the intervention arm (RH 0.61, 95%CI: 0.41–0.91, P = 0.02). Interpretation There was no difference by study arm in the rate of acquisition of CT or GC. However, our per-protocol results suggest that consistent use of the diaphragm may reduce acquisition of GC. Trial

  10. Performance Characteristics of the QUANTIPLEX HIV-1 RNA 3.0 Assay for Detection and Quantitation of Human Immunodeficiency Virus Type 1 RNA in Plasma

    PubMed Central

    Erice, Alejo; Brambilla, Donald; Bremer, James; Jackson, J. Brooks; Kokka, Robert; Yen-Lieberman, Belinda; Coombs, Robert W.

    2000-01-01

    The QUANTIPLEX HIV-1 RNA assay, version 3.0 (a branched DNA, version 3.0, assay [bDNA 3.0 assay]), was evaluated by analyzing spiked and clinical plasma samples and was compared with the AMPLICOR HIV-1 MONITOR Ultrasensitive (ultrasensitive reverse transcription-PCR [US-RT-PCR]) method. A panel of spiked plasma samples that contained 0 to 750,000 copies of human immunodeficiency virus type 1 (HIV-1) RNA per ml was tested four times in each of four laboratories (1,344 assays). Negative results (<50 copies/ml) were obtained in 30 of 32 (94%) assays with seronegative samples, 66 of 128 (52%) assays with HIV-1 RNA at 50 copies/ml, and 5 of 128 (4%) assays with HIV-1 RNA at 100 copies/ml. The assay was linear from 100 to 500,000 copies/ml. The within-run standard deviation (SD) of the log10 estimated HIV-1 RNA concentration was 0.08 at 1,000 to 500,000 copies/ml, increased below 1,000 copies/ml, and was 0.17 at 100 copies/ml. Between-run reproducibility at 100 to 500 copies/ml was <0.10 log10 in most comparisons. Interlaboratory differences across runs were ≤0.10 log10 at all concentrations examined. A subset of the panel (25 to 500 copies/ml) was also analyzed by the US-RT-PCR assay. The within-run SD varied inversely with the log10 HIV-1 RNA concentration but was higher than the SD for the bDNA 3.0 assay at all concentrations. Log-log regression analysis indicated that the two methods produced very similar estimates at 100 to 500 copies/ml. In parallel testing of clinical specimens with low HIV-1 RNA levels, 80 plasma samples with <50 copies/ml by the US-RT-PCR assay had <50 copies/ml when they were retested by the bDNA 3.0 assay. In contrast, 11 of 78 (14%) plasma samples with <50 copies/ml by the bDNA 3.0 assay had ≥50 copies/ml when they were retested by the US-RT-PCR assay (median, 86 copies/ml; range, 50 to 217 copies/ml). Estimation of bDNA 3.0 values of <50 copies/ml by extending the standard curve of the assay showed that these samples with discrepant