Sample records for acceptable image quality

  1. Quantitative analysis of image quality for acceptance and commissioning of an MRI simulator with a semiautomatic method.

    PubMed

    Chen, Xinyuan; Dai, Jianrong

    2018-05-01

    Magnetic Resonance Imaging (MRI) simulation differs from diagnostic MRI in purpose, technical requirements, and implementation. We propose a semiautomatic method for image acceptance and commissioning for the scanner, the radiofrequency (RF) coils, and pulse sequences for an MRI simulator. The ACR MRI accreditation large phantom was used for image quality analysis with seven parameters. Standard ACR sequences with a split head coil were adopted to examine the scanner's basic performance. The performance of simulation RF coils were measured and compared using the standard sequence with different clinical diagnostic coils. We used simulation sequences with simulation coils to test the quality of image and advanced performance of the scanner. Codes and procedures were developed for semiautomatic image quality analysis. When using standard ACR sequences with a split head coil, image quality passed all ACR recommended criteria. The image intensity uniformity with a simulation RF coil decreased about 34% compared with the eight-channel diagnostic head coil, while the other six image quality parameters were acceptable. Those two image quality parameters could be improved to more than 85% by built-in intensity calibration methods. In the simulation sequences test, the contrast resolution was sensitive to the FOV and matrix settings. The geometric distortion of simulation sequences such as T1-weighted and T2-weighted images was well-controlled in the isocenter and 10 cm off-center within a range of ±1% (2 mm). We developed a semiautomatic image quality analysis method for quantitative evaluation of images and commissioning of an MRI simulator. The baseline performances of simulation RF coils and pulse sequences have been established for routine QA. © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  2. Survey of the prevalence and methodology of quality assurance for B-mode ultrasound image quality among veterinary sonographers.

    PubMed

    Hoscheit, Larry P; Heng, Hock Gan; Lim, Chee Kin; Weng, Hsin-Yi

    2018-05-01

    Image quality in B-mode ultrasound is important as it reflects the diagnostic accuracy and diagnostic information provided during clinical scanning. Quality assurance programs for B-mode ultrasound systems/components are comprised of initial quality acceptance testing and subsequent regularly scheduled quality control testing. The importance of quality assurance programs for B-mode ultrasound image quality using ultrasound phantoms is well documented in the human medical and medical physics literature. The purpose of this prospective, cross-sectional, survey study was to determine the prevalence and methodology of quality acceptance testing and quality control testing of image quality for ultrasound system/components among veterinary sonographers. An online electronic survey was sent to 1497 members of veterinary imaging organizations: the American College of Veterinary Radiology, the Veterinary Ultrasound Society, and the European Association of Veterinary Diagnostic Imaging, and a total of 167 responses were received. The results showed that the percentages of veterinary sonographers performing quality acceptance testing and quality control testing are 42% (64/151; 95% confidence interval 34-52%) and 26% (40/156: 95% confidence interval 19-33%) respectively. Of the respondents who claimed to have quality acceptance testing or quality control testing of image quality in place for their ultrasound system/components, 0% have performed quality acceptance testing or quality control testing correctly (quality acceptance testing 95% confidence interval: 0-6%, quality control testing 95% confidence interval: 0-11%). Further education and guidelines are recommended for veterinary sonographers in the area of quality acceptance testing and quality control testing for B-mode ultrasound equipment/components. © 2018 American College of Veterinary Radiology.

  3. An image assessment study of image acceptability of the Galileo low gain antenna mission

    NASA Technical Reports Server (NTRS)

    Chuang, S. L.; Haines, R. F.; Grant, T.; Gold, Yaron; Cheung, Kar-Ming

    1994-01-01

    This paper describes a study conducted by NASA Ames Research Center (ARC) in collaboration with the Jet Propulsion Laboratory (JPL), Pasadena, California on the image acceptability of the Galileo Low Gain Antenna mission. The primary objective of the study is to determine the impact of the Integer Cosine Transform (ICT) compression algorithm on Galilean images of atmospheric bodies, moons, asteroids and Jupiter's rings. The approach involved fifteen volunteer subjects representing twelve institutions involved with the Galileo Solid State Imaging (SSI) experiment. Four different experiment specific quantization tables (q-table) and various compression stepsizes (q-factor) to achieve different compression ratios were used. It then determined the acceptability of the compressed monochromatic astronomical images as evaluated by Galileo SSI mission scientists. Fourteen different images were evaluated. Each observer viewed two versions of the same image side by side on a high resolution monitor, each was compressed using a different quantization stepsize. They were requested to select which image had the highest overall quality to support them in carrying out their visual evaluations of image content. Then they rated both images using a scale from one to five on its judged degree of usefulness. Up to four pre-selected types of images were presented with and without noise to each subject based upon results of a previously administered survey of their image preferences. Fourteen different images in seven image groups were studied. The results showed that: (1) acceptable compression ratios vary widely with the type of images; (2) noisy images detract greatly from image acceptability and acceptable compression ratios; and (3) atmospheric images of Jupiter seem to have higher compression ratios of 4 to 5 times that of some clear surface satellite images.

  4. Scanning technology selection impacts acceptability and usefulness of image-rich content.

    PubMed

    Alpi, Kristine M; Brown, James C; Neel, Jennifer A; Grindem, Carol B; Linder, Keith E; Harper, James B

    2016-01-01

    Clinical and research usefulness of articles can depend on image quality. This study addressed whether scans of figures in black and white (B&W), grayscale, or color, or portable document format (PDF) to tagged image file format (TIFF) conversions as provided by interlibrary loan or document delivery were viewed as acceptable or useful by radiologists or pathologists. Residency coordinators selected eighteen figures from studies from radiology, clinical pathology, and anatomic pathology journals. With original PDF controls, each figure was prepared in three or four experimental conditions: PDF conversion to TIFF, and scans from print in B&W, grayscale, and color. Twelve independent observers indicated whether they could identify the features and whether the image quality was acceptable. They also ranked all the experimental conditions of each figure in terms of usefulness. Of 982 assessments of 87 anatomic pathology, 83 clinical pathology, and 77 radiology images, 471 (48%) were unidentifiable. Unidentifiability of originals (4%) and conversions (10%) was low. For scans, unidentifiability ranged from 53% for color, to 74% for grayscale, to 97% for B&W. Of 987 responses about acceptability (n=405), 41% were said to be unacceptable, 97% of B&W, 66% of grayscale, 41% of color, and 1% of conversions. Hypothesized order (original, conversion, color, grayscale, B&W) matched 67% of rankings (n=215). PDF to TIFF conversion provided acceptable content. Color images are rarely useful in grayscale (12%) or B&W (less than 1%). Acceptability of grayscale scans of noncolor originals was 52%. Digital originals are needed for most images. Print images in color or grayscale should be scanned using those modalities.

  5. Scanning technology selection impacts acceptability and usefulness of image-rich content*†

    PubMed Central

    Alpi, Kristine M.; Brown, James C.; Neel, Jennifer A.; Grindem, Carol B.; Linder, Keith E.; Harper, James B.

    2016-01-01

    Objective Clinical and research usefulness of articles can depend on image quality. This study addressed whether scans of figures in black and white (B&W), grayscale, or color, or portable document format (PDF) to tagged image file format (TIFF) conversions as provided by interlibrary loan or document delivery were viewed as acceptable or useful by radiologists or pathologists. Methods Residency coordinators selected eighteen figures from studies from radiology, clinical pathology, and anatomic pathology journals. With original PDF controls, each figure was prepared in three or four experimental conditions: PDF conversion to TIFF, and scans from print in B&W, grayscale, and color. Twelve independent observers indicated whether they could identify the features and whether the image quality was acceptable. They also ranked all the experimental conditions of each figure in terms of usefulness. Results Of 982 assessments of 87 anatomic pathology, 83 clinical pathology, and 77 radiology images, 471 (48%) were unidentifiable. Unidentifiability of originals (4%) and conversions (10%) was low. For scans, unidentifiability ranged from 53% for color, to 74% for grayscale, to 97% for B&W. Of 987 responses about acceptability (n=405), 41% were said to be unacceptable, 97% of B&W, 66% of grayscale, 41% of color, and 1% of conversions. Hypothesized order (original, conversion, color, grayscale, B&W) matched 67% of rankings (n=215). Conclusions PDF to TIFF conversion provided acceptable content. Color images are rarely useful in grayscale (12%) or B&W (less than 1%). Acceptability of grayscale scans of noncolor originals was 52%. Digital originals are needed for most images. Print images in color or grayscale should be scanned using those modalities. PMID:26807048

  6. Optimization of image quality and dose for Varian aS500 electronic portal imaging devices (EPIDs).

    PubMed

    McGarry, C K; Grattan, M W D; Cosgrove, V P

    2007-12-07

    This study was carried out to investigate whether the electronic portal imaging (EPI) acquisition process could be optimized, and as a result tolerance and action levels be set for the PIPSPro QC-3V phantom image quality assessment. The aim of the optimization process was to reduce the dose delivered to the patient while maintaining a clinically acceptable image quality. This is of interest when images are acquired in addition to the planned patient treatment, rather than images being acquired using the treatment field during a patient's treatment. A series of phantoms were used to assess image quality for different acquisition settings relative to the baseline values obtained following acceptance testing. Eight Varian aS500 EPID systems on four matched Varian 600C/D linacs and four matched Varian 2100C/D linacs were compared for consistency of performance and images were acquired at the four main orthogonal gantry angles. Images were acquired using a 6 MV beam operating at 100 MU min(-1) and the low-dose acquisition mode. Doses used in the comparison were measured using a Farmer ionization chamber placed at d(max) in solid water. The results demonstrated that the number of reset frames did not have any influence on the image contrast, but the number of frame averages did. The expected increase in noise with corresponding decrease in contrast was also observed when reducing the number of frame averages. The optimal settings for the low-dose acquisition mode with respect to image quality and dose were found to be one reset frame and three frame averages. All patients at the Northern Ireland Cancer Centre are now imaged using one reset frame and three frame averages in the 6 MV 100 MU min(-1) low-dose acquisition mode. Routine EPID QC contrast tolerance (+/-10) and action (+/-20) levels using the PIPSPro phantom based around expected values of 190 (Varian 600C/D) and 225 (Varian 2100C/D) have been introduced. The dose at dmax from electronic portal imaging has been

  7. Automated Quality Assessment of Colour Fundus Images for Diabetic Retinopathy Screening in Telemedicine.

    PubMed

    Saha, Sajib Kumar; Fernando, Basura; Cuadros, Jorge; Xiao, Di; Kanagasingam, Yogesan

    2018-04-27

    Fundus images obtained in a telemedicine program are acquired at different sites that are captured by people who have varying levels of experience. These result in a relatively high percentage of images which are later marked as unreadable by graders. Unreadable images require a recapture which is time and cost intensive. An automated method that determines the image quality during acquisition is an effective alternative. To determine the image quality during acquisition, we describe here an automated method for the assessment of image quality in the context of diabetic retinopathy. The method explicitly applies machine learning techniques to access the image and to determine 'accept' and 'reject' categories. 'Reject' category image requires a recapture. A deep convolution neural network is trained to grade the images automatically. A large representative set of 7000 colour fundus images was used for the experiment which was obtained from the EyePACS that were made available by the California Healthcare Foundation. Three retinal image analysis experts were employed to categorise these images into 'accept' and 'reject' classes based on the precise definition of image quality in the context of DR. The network was trained using 3428 images. The method shows an accuracy of 100% to successfully categorise 'accept' and 'reject' images, which is about 2% higher than the traditional machine learning method. On a clinical trial, the proposed method shows 97% agreement with human grader. The method can be easily incorporated with the fundus image capturing system in the acquisition centre and can guide the photographer whether a recapture is necessary or not.

  8. Implant image quality in dental radiographs recorded using a customized imaging guide or a standard film holder.

    PubMed

    Schropp, Lars; Stavropoulos, Andreas; Spin-Neto, Rubens; Wenzel, Ann

    2012-01-01

    To compare a customized imaging guide and a standard film holder for obtaining optimally projected intraoral radiographs of dental implants. Intraoral radiographs of four screw-type implants with different inclination placed in an upper or lower dental phantom model were recorded by 32 groups of examiners after a short instruction in the use of the RB-RB/LB-LB mnemonic rule. Half of the examiners recorded the images using a standard film holder and the other half used a customized imaging guide. Each radiograph was assessed under blinded conditions with regard to rendering of the implant threads and was assigned to one of four quality categories: (1) perfect, (2) not perfect, but clinically acceptable, (3) not acceptable, and (4) hopeless. For the upper jaw, the same number of exposures per implant were made to achieve an acceptable image (P=0.86) by the standard film holder method (median=2) and the imaging guide method (median=2). For the lower jaw, medians for the imaging guide method and the film holder method were 1 and 2, respectively (P=0.004). For the imaging guide method, the first exposure was rated as perfect/acceptable in 62% of the cases and for the film holder method in 41% of the cases (P=0.013). After ≤ 2 exposures, 78% (imaging guide method) and 69% (film holder method) of the implant images were perfect/acceptable (P=0.23). The implant inclination did not have a major influence on the outcomes. Perfect or acceptable images were achieved after two exposures with the same frequency either using a customized imaging guide method or a standard film holder method. However, the use of a customized imaging guide method was overall significantly superior to a standard film holder method in terms of obtaining perfect or acceptable images with only one exposure. © 2011 John Wiley & Sons A/S.

  9. The effects of video compression on acceptability of images for monitoring life sciences experiments

    NASA Astrophysics Data System (ADS)

    Haines, Richard F.; Chuang, Sherry L.

    1992-07-01

    Future manned space operations for Space Station Freedom will call for a variety of carefully planned multimedia digital communications, including full-frame-rate color video, to support remote operations of scientific experiments. This paper presents the results of an investigation to determine if video compression is a viable solution to transmission bandwidth constraints. It reports on the impact of different levels of compression and associated calculational parameters on image acceptability to investigators in life-sciences research at ARC. Three nonhuman life-sciences disciplines (plant, rodent, and primate biology) were selected for this study. A total of 33 subjects viewed experimental scenes in their own scientific disciplines. Ten plant scientists viewed still images of wheat stalks at various stages of growth. Each image was compressed to four different compression levels using the Joint Photographic Expert Group (JPEG) standard algorithm, and the images were presented in random order. Twelve and eleven staffmembers viewed 30-sec videotaped segments showing small rodents and a small primate, respectively. Each segment was repeated at four different compression levels in random order using an inverse cosine transform (ICT) algorithm. Each viewer made a series of subjective image-quality ratings. There was a significant difference in image ratings according to the type of scene viewed within disciplines; thus, ratings were scene dependent. Image (still and motion) acceptability does, in fact, vary according to compression level. The JPEG still-image-compression levels, even with the large range of 5:1 to 120:1 in this study, yielded equally high levels of acceptability. In contrast, the ICT algorithm for motion compression yielded a sharp decline in acceptability below 768 kb/sec. Therefore, if video compression is to be used as a solution for overcoming transmission bandwidth constraints, the effective management of the ratio and compression parameters

  10. The effects of video compression on acceptability of images for monitoring life sciences experiments

    NASA Technical Reports Server (NTRS)

    Haines, Richard F.; Chuang, Sherry L.

    1992-01-01

    Future manned space operations for Space Station Freedom will call for a variety of carefully planned multimedia digital communications, including full-frame-rate color video, to support remote operations of scientific experiments. This paper presents the results of an investigation to determine if video compression is a viable solution to transmission bandwidth constraints. It reports on the impact of different levels of compression and associated calculational parameters on image acceptability to investigators in life-sciences research at ARC. Three nonhuman life-sciences disciplines (plant, rodent, and primate biology) were selected for this study. A total of 33 subjects viewed experimental scenes in their own scientific disciplines. Ten plant scientists viewed still images of wheat stalks at various stages of growth. Each image was compressed to four different compression levels using the Joint Photographic Expert Group (JPEG) standard algorithm, and the images were presented in random order. Twelve and eleven staffmembers viewed 30-sec videotaped segments showing small rodents and a small primate, respectively. Each segment was repeated at four different compression levels in random order using an inverse cosine transform (ICT) algorithm. Each viewer made a series of subjective image-quality ratings. There was a significant difference in image ratings according to the type of scene viewed within disciplines; thus, ratings were scene dependent. Image (still and motion) acceptability does, in fact, vary according to compression level. The JPEG still-image-compression levels, even with the large range of 5:1 to 120:1 in this study, yielded equally high levels of acceptability. In contrast, the ICT algorithm for motion compression yielded a sharp decline in acceptability below 768 kb/sec. Therefore, if video compression is to be used as a solution for overcoming transmission bandwidth constraints, the effective management of the ratio and compression parameters

  11. Histogram Matching Extends Acceptable Signal Strength Range on Optical Coherence Tomography Images

    PubMed Central

    Chen, Chieh-Li; Ishikawa, Hiroshi; Wollstein, Gadi; Bilonick, Richard A.; Sigal, Ian A.; Kagemann, Larry; Schuman, Joel S.

    2015-01-01

    Purpose. We minimized the influence of image quality variability, as measured by signal strength (SS), on optical coherence tomography (OCT) thickness measurements using the histogram matching (HM) method. Methods. We scanned 12 eyes from 12 healthy subjects with the Cirrus HD-OCT device to obtain a series of OCT images with a wide range of SS (maximal range, 1–10) at the same visit. For each eye, the histogram of an image with the highest SS (best image quality) was set as the reference. We applied HM to the images with lower SS by shaping the input histogram into the reference histogram. Retinal nerve fiber layer (RNFL) thickness was automatically measured before and after HM processing (defined as original and HM measurements), and compared to the device output (device measurements). Nonlinear mixed effects models were used to analyze the relationship between RNFL thickness and SS. In addition, the lowest tolerable SSs, which gave the RNFL thickness within the variability margin of manufacturer recommended SS range (6–10), were determined for device, original, and HM measurements. Results. The HM measurements showed less variability across a wide range of image quality than the original and device measurements (slope = 1.17 vs. 4.89 and 1.72 μm/SS, respectively). The lowest tolerable SS was successfully reduced to 4.5 after HM processing. Conclusions. The HM method successfully extended the acceptable SS range on OCT images. This would qualify more OCT images with low SS for clinical assessment, broadening the OCT application to a wider range of subjects. PMID:26066749

  12. Application of QC_DR software for acceptance testing and routine quality control of direct digital radiography systems: initial experiences using the Italian Association of Physicist in Medicine quality control protocol.

    PubMed

    Nitrosi, Andrea; Bertolini, Marco; Borasi, Giovanni; Botti, Andrea; Barani, Adriana; Rivetti, Stefano; Pierotti, Luisa

    2009-12-01

    Ideally, medical x-ray imaging systems should be designed to deliver maximum image quality at an acceptable radiation risk to the patient. Quality assurance procedures are employed to ensure that these standards are maintained. A quality control protocol for direct digital radiography (DDR) systems is described and discussed. Software to automatically process and analyze the required images was developed. In this paper, the initial results obtained on equipment of different DDR manufacturers were reported. The protocol was developed to highlight even small discrepancies in standard operating performance.

  13. Light on body image treatment: acceptance through mindfulness.

    PubMed

    Stewart, Tiffany M

    2004-11-01

    The treatment of body image has to be multifaceted and should be directed toward the treatment of the whole individual-body, mind, and spirit-with an ultimate culmination of acceptance and compassion for the self. This article presents information on a mindful approach to the treatment of body image as it pertains to concerns with body size and shape. This approach fosters the idea that the treatment process should be one of observation, nonjudgment, neutrality, and acceptance. To this end, this article will depict the conceptualization of body image treatment from a mindful perspective, in which mindfulness serves as the foundation on which the multiple facets of treatment are built. The core components of body image treatment (i.e., cognitive, perceptual, behavioral, and emotional), in the context of mindfulness, are discussed as they relate to the treatment of body image disturbance. This article may be viewed as a theoretical overview of a new treatment concept for body image disturbance.

  14. How much image noise can be added in cardiac x-ray imaging without loss in perceived image quality?

    NASA Astrophysics Data System (ADS)

    Gislason-Lee, Amber J.; Kumcu, Asli; Kengyelics, Stephen M.; Rhodes, Laura A.; Davies, Andrew G.

    2015-03-01

    Dynamic X-ray imaging systems are used for interventional cardiac procedures to treat coronary heart disease. X-ray settings are controlled automatically by specially-designed X-ray dose control mechanisms whose role is to ensure an adequate level of image quality is maintained with an acceptable radiation dose to the patient. Current commonplace dose control designs quantify image quality by performing a simple technical measurement directly from the image. However, the utility of cardiac X-ray images is in their interpretation by a cardiologist during an interventional procedure, rather than in a technical measurement. With the long term goal of devising a clinically-relevant image quality metric for an intelligent dose control system, we aim to investigate the relationship of image noise with clinical professionals' perception of dynamic image sequences. Computer-generated noise was added, in incremental amounts, to angiograms of five different patients selected to represent the range of adult cardiac patient sizes. A two alternative forced choice staircase experiment was used to determine the amount of noise which can be added to a patient image sequences without changing image quality as perceived by clinical professionals. Twenty-five viewing sessions (five for each patient) were completed by thirteen observers. Results demonstrated scope to increase the noise of cardiac X-ray images by up to 21% +/- 8% before it is noticeable by clinical professionals. This indicates a potential for 21% radiation dose reduction since X-ray image noise and radiation dose are directly related; this would be beneficial to both patients and personnel.

  15. Sexual Function Is Correlated With Body Image and Partnership Quality in Female University Students.

    PubMed

    Wallwiener, Stephanie; Strohmaier, Jana; Wallwiener, Lisa-Maria; Schönfisch, Birgitt; Zipfel, Stephan; Brucker, Sara Y; Rietschel, Marcella; Wallwiener, Christian W

    2016-10-01

    According to the World Health Organization definition, sexual health is more than mere physical sexual function; it also encompasses emotional, mental, and social well-being in relation to sexuality and is not merely the absence of dysfunction or disease. In line with this definition, various studies have reported that female sexual function is associated with partnership quality, body image, and body self-acceptance. To investigate whether female sexual function is influenced by (i) body self-acceptance and (ii) partnership quality, as important factors in psychosocial well-being, and (iii) whether the effects of body self-acceptance are moderated by partnership quality. In total, 2,685 female medical students no older than 35 years from Germany, Austria, and Switzerland completed an anonymous online questionnaire comprising the Female Sexual Function Index (FSFI) and the Self-Acceptance of the Body Scale. Respondents were asked to state whether they had been in a steady partnership in the preceding 6 months. When present, the quality of the partnership status was rated (enamoredness, love, friendship, or conflicted). To determine correlations, group differences, and moderating effects among body self-acceptance, partnership quality, and sexual function, the data were analyzed using Spearman correlations, Kruskal-Wallis tests, and analyses of variance. Female sexual function (FSFI total score). (i) In sexually active women, higher FSFI scores were significantly associated with greater body self-acceptance and a steady partnership during the preceding 6 months. (ii) Total FSFI scores were highest in women who described their partnership as enamored (29.45) or loving (28.55). Lower scores were observed in single women (26.71) and in women who described their partnerships as friendship (25.76) or as emotionally conflicted (23.41). (iii) Total FSFI score was affected by an interaction between body self-acceptance and partnership quality. Body self- acceptance was

  16. Development and validation of a visual grading scale for assessing image quality of AP pelvis radiographic images.

    PubMed

    Mraity, Hussien A A B; England, Andrew; Cassidy, Simon; Eachus, Peter; Dominguez, Alejandro; Hogg, Peter

    2016-01-01

    The aim of this article was to apply psychometric theory to develop and validate a visual grading scale for assessing the visual perception of digital image quality anteroposterior (AP) pelvis. Psychometric theory was used to guide scale development. Seven phantom and seven cadaver images of visually and objectively predetermined quality were used to help assess scale reliability and validity. 151 volunteers scored phantom images, and 184 volunteers scored cadaver images. Factor analysis and Cronbach's alpha were used to assess scale validity and reliability. A 24-item scale was produced. Aggregated mean volunteer scores for each image correlated with the rank order of the visually and objectively predetermined image qualities. Scale items had good interitem correlation (≥0.2) and high factor loadings (≥0.3). Cronbach's alpha (reliability) revealed that the scale has acceptable levels of internal reliability for both phantom and cadaver images (α = 0.8 and 0.9, respectively). Factor analysis suggested that the scale is multidimensional (assessing multiple quality themes). This study represents the first full development and validation of a visual image quality scale using psychometric theory. It is likely that this scale will have clinical, training and research applications. This article presents data to create and validate visual grading scales for radiographic examinations. The visual grading scale, for AP pelvis examinations, can act as a validated tool for future research, teaching and clinical evaluations of image quality.

  17. Development and validation of a visual grading scale for assessing image quality of AP pelvis radiographic images

    PubMed Central

    England, Andrew; Cassidy, Simon; Eachus, Peter; Dominguez, Alejandro; Hogg, Peter

    2016-01-01

    Objective: The aim of this article was to apply psychometric theory to develop and validate a visual grading scale for assessing the visual perception of digital image quality anteroposterior (AP) pelvis. Methods: Psychometric theory was used to guide scale development. Seven phantom and seven cadaver images of visually and objectively predetermined quality were used to help assess scale reliability and validity. 151 volunteers scored phantom images, and 184 volunteers scored cadaver images. Factor analysis and Cronbach's alpha were used to assess scale validity and reliability. Results: A 24-item scale was produced. Aggregated mean volunteer scores for each image correlated with the rank order of the visually and objectively predetermined image qualities. Scale items had good interitem correlation (≥0.2) and high factor loadings (≥0.3). Cronbach's alpha (reliability) revealed that the scale has acceptable levels of internal reliability for both phantom and cadaver images (α = 0.8 and 0.9, respectively). Factor analysis suggested that the scale is multidimensional (assessing multiple quality themes). Conclusion: This study represents the first full development and validation of a visual image quality scale using psychometric theory. It is likely that this scale will have clinical, training and research applications. Advances in knowledge: This article presents data to create and validate visual grading scales for radiographic examinations. The visual grading scale, for AP pelvis examinations, can act as a validated tool for future research, teaching and clinical evaluations of image quality. PMID:26943836

  18. Paediatric x-ray radiation dose reduction and image quality analysis.

    PubMed

    Martin, L; Ruddlesden, R; Makepeace, C; Robinson, L; Mistry, T; Starritt, H

    2013-09-01

    Collaboration of multiple staff groups has resulted in significant reduction in the risk of radiation-induced cancer from radiographic x-ray exposure during childhood. In this study at an acute NHS hospital trust, a preliminary audit identified initial exposure factors. These were compared with European and UK guidance, leading to the introduction of new factors that were in compliance with European guidance on x-ray tube potentials. Image quality was assessed using standard anatomical criteria scoring, and visual grading characteristics analysis assessed the impact on image quality of changes in exposure factors. This analysis determined the acceptability of gradual radiation dose reduction below the European and UK guidance levels. Chest and pelvis exposures were optimised, achieving dose reduction for each age group, with 7%-55% decrease in critical organ dose. Clinicians confirmed diagnostic image quality throughout the iterative process. Analysis of images acquired with preliminary and final exposure factors indicated an average visual grading analysis result of 0.5, demonstrating equivalent image quality. The optimisation process and final radiation doses are reported for Carestream computed radiography to aid other hospitals in minimising radiation risks to children.

  19. Light on Body Image Treatment: Acceptance Through Mindfulness

    ERIC Educational Resources Information Center

    Stewart, Tiffany M.

    2004-01-01

    The treatment of body image has to be multifaceted and should be directed toward the treatment of the whole individual - body, mind, and spirit - with an ultimate culmination of acceptance and compassion for the self. This article presents information on a mindful approach to the treatment of body image as it pertains to concerns with body size…

  20. Social image quality

    NASA Astrophysics Data System (ADS)

    Qiu, Guoping; Kheiri, Ahmed

    2011-01-01

    Current subjective image quality assessments have been developed in the laboratory environments, under controlledconditions, and are dependent on the participation of limited numbers of observers. In this research, with the help of Web 2.0 and social media technology, a new method for building a subjective image quality metric has been developed where the observers are the Internet users. A website with a simple user interface that enables Internet users from anywhere at any time to vote for a better quality version of a pair of the same image has been constructed. Users' votes are recorded and used to rank the images according to their perceived visual qualities. We have developed three rank aggregation algorithms to process the recorded pair comparison data, the first uses a naive approach, the second employs a Condorcet method, and the third uses the Dykstra's extension of Bradley-Terry method. The website has been collecting data for about three months and has accumulated over 10,000 votes at the time of writing this paper. Results show that the Internet and its allied technologies such as crowdsourcing offer a promising new paradigm for image and video quality assessment where hundreds of thousands of Internet users can contribute to building more robust image quality metrics. We have made Internet user generated social image quality (SIQ) data of a public image database available online (http://www.hdri.cs.nott.ac.uk/siq/) to provide the image quality research community with a new source of ground truth data. The website continues to collect votes and will include more public image databases and will also be extended to include videos to collect social video quality (SVQ) data. All data will be public available on the website in due course.

  1. AIR CLEANING FOR ACCEPTABLE INDOOR AIR QUALITY

    EPA Science Inventory

    The paper discusses air cleaning for acceptable indoor air quality. ir cleaning has performed an important role in heating, ventilation, and air-conditioning systems for many years. raditionally, general ventilation air-filtration equipment has been used to protect cooling coils ...

  2. Sci-Thur AM: YIS – 08: Automated Imaging Quality Assurance for Image-Guided Small Animal Irradiators

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Johnstone, Chris; Bazalova-Carter, Magdalena

    Purpose: To develop quality assurance (QA) standards and tolerance levels for image quality of small animal irradiators. Methods: A fully automated in-house QA software for image analysis of a commercial microCT phantom was created. Quantitative analyses of CT linearity, signal-to-noise ratio (SNR), uniformity and noise, geometric accuracy, modulation transfer function (MTF), and CT number evaluation was performed. Phantom microCT scans from seven institutions acquired with varying parameters (kVp, mA, time, voxel size, and frame rate) and five irradiator units (Xstrahl SARRP, PXI X-RAD 225Cx, PXI X-RAD SmART, GE explore CT/RT 140, and GE Explore CT 120) were analyzed. Multi-institutional datamore » sets were compared using our in-house software to establish pass/fail criteria for each QA test. Results: CT linearity (R2>0.996) was excellent at all but Institution 2. Acceptable SNR (>35) and noise levels (<55HU) were obtained at four of the seven institutions, where failing scans were acquired with less than 120mAs. Acceptable MTF (>1.5 lp/mm for MTF=0.2) was obtained at all but Institution 6 due to the largest scan voxel size (0.35mm). The geometric accuracy passed (<1.5%) at five of the seven institutions. Conclusion: Our QA software can be used to rapidly perform quantitative imaging QA for small animal irradiators, accumulate results over time, and display possible changes in imaging functionality from its original performance and/or from the recommended tolerance levels. This tool will aid researchers in maintaining high image quality, enabling precise conformal dose delivery to small animals.« less

  3. Image Quality Assessment of JPEG Compressed Mars Science Laboratory Mastcam Images using Convolutional Neural Networks

    NASA Astrophysics Data System (ADS)

    Kerner, H. R.; Bell, J. F., III; Ben Amor, H.

    2017-12-01

    The Mastcam color imaging system on the Mars Science Laboratory Curiosity rover acquires images within Gale crater for a variety of geologic and atmospheric studies. Images are often JPEG compressed before being downlinked to Earth. While critical for transmitting images on a low-bandwidth connection, this compression can result in image artifacts most noticeable as anomalous brightness or color changes within or near JPEG compression block boundaries. In images with significant high-frequency detail (e.g., in regions showing fine layering or lamination in sedimentary rocks), the image might need to be re-transmitted losslessly to enable accurate scientific interpretation of the data. The process of identifying which images have been adversely affected by compression artifacts is performed manually by the Mastcam science team, costing significant expert human time. To streamline the tedious process of identifying which images might need to be re-transmitted, we present an input-efficient neural network solution for predicting the perceived quality of a compressed Mastcam image. Most neural network solutions require large amounts of hand-labeled training data for the model to learn the target mapping between input (e.g. distorted images) and output (e.g. quality assessment). We propose an automatic labeling method using joint entropy between a compressed and uncompressed image to avoid the need for domain experts to label thousands of training examples by hand. We use automatically labeled data to train a convolutional neural network to estimate the probability that a Mastcam user would find the quality of a given compressed image acceptable for science analysis. We tested our model on a variety of Mastcam images and found that the proposed method correlates well with image quality perception by science team members. When assisted by our proposed method, we estimate that a Mastcam investigator could reduce the time spent reviewing images by a minimum of 70%.

  4. Effects of AEC chamber selection on patient dose and image quality.

    PubMed

    Hawking, Nancy; Elmore, Angie

    2009-01-01

    To determine whether manipulation of the standard automatic exposure control (AEC) chamber selections reduces the patient's entrance skin exposure (ESE) without compromising image quality. Data for density and radiation dose were gathered at 2 clinical locations by exposing abdomen and pelvis phantoms to radiation using 3 AEC chamber selection configurations. ESE (skin dose) was measured using a multipurpose dosimeter. The experiment included both film-screen and computed radiography (CR) systems. For both phantoms, using the 2 outside chambers resulted in the lowest dose on the film-screen and CR systems. In general, optical density (OD) and exposure indicator (EI) remained within acceptable ranges and image quality was maintained using this chamber configuration. Using only the center chamber resulted in the highest dose increases and lowest image quality for film-screen and CR systems. When performing anteroposterior (AP) abdomen and AP pelvis examinations, radiographers can reduce patients' ESE and maintain image quality by selecting the 2 outside AEC chambers. Further research on AEC chamber selection should be conducted for additional anatomical regions.

  5. Image quality improvement in optical coherence tomography using Lucy-Richardson deconvolution algorithm.

    PubMed

    Hojjatoleslami, S A; Avanaki, M R N; Podoleanu, A Gh

    2013-08-10

    Optical coherence tomography (OCT) has the potential for skin tissue characterization due to its high axial and transverse resolution and its acceptable depth penetration. In practice, OCT cannot reach the theoretical resolutions due to imperfections of some of the components used. One way to improve the quality of the images is to estimate the point spread function (PSF) of the OCT system and deconvolve it from the output images. In this paper, we investigate the use of solid phantoms to estimate the PSF of the imaging system. We then utilize iterative Lucy-Richardson deconvolution algorithm to improve the quality of the images. The performance of the proposed algorithm is demonstrated on OCT images acquired from a variety of samples, such as epoxy-resin phantoms, fingertip skin and basaloid larynx and eyelid tissues.

  6. Validation of an image-based technique to assess the perceptual quality of clinical chest radiographs with an observer study

    NASA Astrophysics Data System (ADS)

    Lin, Yuan; Choudhury, Kingshuk R.; McAdams, H. Page; Foos, David H.; Samei, Ehsan

    2014-03-01

    We previously proposed a novel image-based quality assessment technique1 to assess the perceptual quality of clinical chest radiographs. In this paper, an observer study was designed and conducted to systematically validate this technique. Ten metrics were involved in the observer study, i.e., lung grey level, lung detail, lung noise, riblung contrast, rib sharpness, mediastinum detail, mediastinum noise, mediastinum alignment, subdiaphragm-lung contrast, and subdiaphragm area. For each metric, three tasks were successively presented to the observers. In each task, six ROI images were randomly presented in a row and observers were asked to rank the images only based on a designated quality and disregard the other qualities. A range slider on the top of the images was used for observers to indicate the acceptable range based on the corresponding perceptual attribute. Five boardcertificated radiologists from Duke participated in this observer study on a DICOM calibrated diagnostic display workstation and under low ambient lighting conditions. The observer data were analyzed in terms of the correlations between the observer ranking orders and the algorithmic ranking orders. Based on the collected acceptable ranges, quality consistency ranges were statistically derived. The observer study showed that, for each metric, the averaged ranking orders of the participated observers were strongly correlated with the algorithmic orders. For the lung grey level, the observer ranking orders completely accorded with the algorithmic ranking orders. The quality consistency ranges derived from this observer study were close to these derived from our previous study. The observer study indicates that the proposed image-based quality assessment technique provides a robust reflection of the perceptual image quality of the clinical chest radiographs. The derived quality consistency ranges can be used to automatically predict the acceptability of a clinical chest radiograph.

  7. Assessment of image quality in x-ray radiography imaging using a small plasma focus device

    NASA Astrophysics Data System (ADS)

    Kanani, A.; Shirani, B.; Jabbari, I.; Mokhtari, J.

    2014-08-01

    This paper offers a comprehensive investigation of image quality parameters for a small plasma focus as a pulsed hard x-ray source for radiography applications. A set of images were captured from some metal objects and electronic circuits using a low energy plasma focus at different voltages of capacitor bank and different pressures of argon gas. The x-ray source focal spot of this device was obtained to be about 0.6 mm using the penumbra imaging method. The image quality was studied by several parameters such as image contrast, line spread function (LSF) and modulation transfer function (MTF). Results showed that the contrast changes by variations in gas pressure. The best contrast was obtained at a pressure of 0.5 mbar and 3.75 kJ stored energy. The results of x-ray dose from the device showed that about 0.6 mGy is sufficient to obtain acceptable images on the film. The measurements of LSF and MTF parameters were carried out by means of a thin stainless steel wire 0.8 mm in diameter and the cut-off frequency was obtained to be about 1.5 cycles/mm.

  8. Acceptance of disease and the quality of life in patients with enteric stoma.

    PubMed

    Szpilewska, Katarzyna; Juzwiszyn, Jan; Bolanowska, Zofia; Bolanowska, Zofia; Milan, Magdalena; Chabowski, Mariusz; Janczak, Dariusz

    2018-02-28

    The main reason of the emergence of enteric stoma is colorectal cancer. Enteric stoma is a serious health, as well as life problem. In Poland, there are about 6,000 stoma surgeries yearly. It changes the functioning of patients, restricts their daily activity and influences their quality of life significantly. Therefore, in the modern treatment process, all spheres of human life and its surroundings are considered. The evaluation of the quality of life and the level of acceptance of the disease enables us to identify the regions in which patients require attention and help, as well as places to which health promotion among patients with a stoma should be directed. The aim of the study was the evaluation of the acceptance of the disease and the quality of life in patients with colostomy. The study conducted between February 2015 and February 2016 included101 patients with enteric stoma of the Provincial Specialist Hospital in Wrocław at Kamieńskiego St. in the departments of General Surgery with the Subdivision of Traumatology and Orthopedic Surgery, Subdivision of Metabolic Surgery, Subdivision of Endocrine Surgery and Oncological Surgery . Two anonymous questionnaires, i.e., health-related quality of life (HRQoL) and acceptance of illness scale (AIS), were used. 60% (61 people) were women, and 40% (40 people) were men. The mean age was 48; the youngest respondent was 20 years old, and the oldest was 79 years old. 17% of the respondents had primary education, 25% - vocational, 33% - secondary and 26% - university-level. The most common reason for the emergence of a stoma among the respondents was colon cancer (44%), followed by: mechanical bowel onstruction (26%), intestinal damage following injuries (25%), inflammatory bowel disease (6%). The quality of life of patients with colostomy was evaluated in view of their health condition, postoperative recovery, everyday limitations, and self-evaluation. The majority of participants pointed to the deterioration of

  9. Comprehensive quality assurance phantom for cardiovascular imaging systems

    NASA Astrophysics Data System (ADS)

    Lin, Pei-Jan P.

    1998-07-01

    With the advent of high heat loading capacity x-ray tubes, high frequency inverter type generators, and the use of spectral shaping filters, the automatic brightness/exposure control (ABC) circuit logic employed in the new generation of angiographic imaging equipment has been significantly reprogrammed. These new angiographic imaging systems are designed to take advantage of the power train capabilities to yield higher contrast images while maintaining, or lower, the patient exposure. Since the emphasis of the imaging system design has been significantly altered, the system performance parameters one is interested and the phantoms employed for the quality assurance must also change in order to properly evaluate the imaging capability of the cardiovascular imaging systems. A quality assurance (QA) phantom has been under development in this institution and was submitted to various interested organizations such as American Association of Physicists in Medicine (AAPM), Society for Cardiac Angiography & Interventions (SCA&I), and National Electrical Manufacturers Association (NEMA) for their review and input. At the same time, in an effort to establish a unified standard phantom design for the cardiac catheterization laboratories (CCL), SCA&I and NEMA have formed a joint work group in early 1997 to develop a suitable phantom. The initial QA phantom design has since been accepted to serve as the base phantom by the SCA&I- NEMA Joint Work Group (JWG) from which a comprehensive QA Phantom is being developed.

  10. Quality assessment of digital X-ray chest images using an anthropomorphic chest phantom

    NASA Astrophysics Data System (ADS)

    Vodovatov, A. V.; Kamishanskaya, I. G.; Drozdov, A. A.; Bernhardsson, C.

    2017-02-01

    The current study is focused on determining the optimal tube voltage for the conventional X-ray digital chest screening examinations, using a visual grading analysis method. Chest images of an anthropomorphic phantom were acquired in posterior-anterior projection on four digital X-ray units with different detector types. X-ray images obtained with an anthropomorphic phantom were accepted by the radiologists as corresponding to a normal human anatomy, hence allowing using phantoms in image quality trials without limitations.

  11. SU-F-207-02: Use of Postmortem Subjects for Subjective Image Quality Assessment in Abdominal CT Protocols with Iterative Reconstruction

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Mench, A; Lipnharski, I; Carranza, C

    Purpose: New radiation dose reduction technologies are emerging constantly in the medical imaging field. The latest of these technologies, iterative reconstruction (IR) in CT, presents the ability to reduce dose significantly and hence provides great opportunity for CT protocol optimization. However, without effective analysis of image quality, the reduction in radiation exposure becomes irrelevant. This work explores the use of postmortem subjects as an image quality assessment medium for protocol optimizations in abdominal CT. Methods: Three female postmortem subjects were scanned using the Abdomen-Pelvis (AP) protocol at reduced minimum tube current and target noise index (SD) settings of 12.5, 17.5,more » 20.0, and 25.0. Images were reconstructed using two strengths of iterative reconstruction. Radiologists and radiology residents from several subspecialties were asked to evaluate 8 AP image sets including the current facility default scan protocol and 7 scans with the parameters varied as listed above. Images were viewed in the soft tissue window and scored on a 3-point scale as acceptable, borderline acceptable, and unacceptable for diagnosis. The facility default AP scan was identified to the reviewer while the 7 remaining AP scans were randomized and de-identified of acquisition and reconstruction details. The observers were also asked to comment on the subjective image quality criteria they used for scoring images. This included visibility of specific anatomical structures and tissue textures. Results: Radiologists scored images as acceptable or borderline acceptable for target noise index settings of up to 20. Due to the postmortem subjects’ close representation of living human anatomy, readers were able to evaluate images as they would those of actual patients. Conclusion: Postmortem subjects have already been proven useful for direct CT organ dose measurements. This work illustrates the validity of their use for the crucial evaluation of image

  12. Decision theory applied to image quality control in radiology.

    PubMed

    Lessa, Patrícia S; Caous, Cristofer A; Arantes, Paula R; Amaro, Edson; de Souza, Fernando M Campello

    2008-11-13

    The present work aims at the application of the decision theory to radiological image quality control (QC) in diagnostic routine. The main problem addressed in the framework of decision theory is to accept or reject a film lot of a radiology service. The probability of each decision of a determined set of variables was obtained from the selected films. Based on a radiology service routine a decision probability function was determined for each considered group of combination characteristics. These characteristics were related to the film quality control. These parameters were also framed in a set of 8 possibilities, resulting in 256 possible decision rules. In order to determine a general utility application function to access the decision risk, we have used a simple unique parameter called r. The payoffs chosen were: diagnostic's result (correct/incorrect), cost (high/low), and patient satisfaction (yes/no) resulting in eight possible combinations. Depending on the value of r, more or less risk will occur related to the decision-making. The utility function was evaluated in order to determine the probability of a decision. The decision was made with patients or administrators' opinions from a radiology service center. The model is a formal quantitative approach to make a decision related to the medical imaging quality, providing an instrument to discriminate what is really necessary to accept or reject a film or a film lot. The method presented herein can help to access the risk level of an incorrect radiological diagnosis decision.

  13. Reproducibility of Mammography Units, Film Processing and Quality Imaging

    NASA Astrophysics Data System (ADS)

    Gaona, Enrique

    2003-09-01

    The purpose of this study was to carry out an exploratory survey of the problems of quality control in mammography and processors units as a diagnosis of the current situation of mammography facilities. Measurements of reproducibility, optical density, optical difference and gamma index are included. Breast cancer is the most frequently diagnosed cancer and is the second leading cause of cancer death among women in the Mexican Republic. Mammography is a radiographic examination specially designed for detecting breast pathology. We found that the problems of reproducibility of AEC are smaller than the problems of processors units because almost all processors fall outside of the acceptable variation limits and they can affect the mammography quality image and the dose to breast. Only four mammography units agree with the minimum score established by ACR and FDA for the phantom image.

  14. Resource quality or competition: why increase resource acceptance in the presence of conspecifics?

    PubMed Central

    Nufio, César R.; Papaj, Daniel R.

    2011-01-01

    Some animal species increase resource acceptance rates in the presence of conspecifics. Such responses may be adaptive if the presence of conspecifics is a reliable indicator of resource quality. Similarly, these responses could represent an adaptive reduction in choosiness under high levels of scramble competition. Although high resource quality and high levels of scramble competition should both favor increased resource acceptance, the contexts in which the increase occurs should differ. In this paper, we tested the effect of social environment on egg-laying and aggressive behavior in the walnut fly, Rhagoletis juglandis, in multiple contexts to determine whether increased resource acceptance in the presence of conspecifics was better viewed as a response to increased host quality or increased competition. We found that grouped females oviposit more readily than isolated females when provided small (low-quality) artificial hosts but not when provided large (high-quality) artificial hosts, indicating that conspecific presence reduces choosiness. Increased resource acceptance was observed even when exposure to conspecifics was temporally or spatially separate from exposure to the resource. Finally, we found that individuals showed reduced aggression after being housed in groups, as expected under high levels of scramble competition. These results indicate that the pattern of resource acceptance in the presence of conspecifics may be better viewed as a response to increased scramble competition rather than as a response to public information about resource quality. PMID:22479135

  15. Resource quality or competition: why increase resource acceptance in the presence of conspecifics?

    PubMed

    Davis, Jeremy M; Nufio, César R; Papaj, Daniel R

    2011-07-01

    Some animal species increase resource acceptance rates in the presence of conspecifics. Such responses may be adaptive if the presence of conspecifics is a reliable indicator of resource quality. Similarly, these responses could represent an adaptive reduction in choosiness under high levels of scramble competition. Although high resource quality and high levels of scramble competition should both favor increased resource acceptance, the contexts in which the increase occurs should differ. In this paper, we tested the effect of social environment on egg-laying and aggressive behavior in the walnut fly, Rhagoletis juglandis, in multiple contexts to determine whether increased resource acceptance in the presence of conspecifics was better viewed as a response to increased host quality or increased competition. We found that grouped females oviposit more readily than isolated females when provided small (low-quality) artificial hosts but not when provided large (high-quality) artificial hosts, indicating that conspecific presence reduces choosiness. Increased resource acceptance was observed even when exposure to conspecifics was temporally or spatially separate from exposure to the resource. Finally, we found that individuals showed reduced aggression after being housed in groups, as expected under high levels of scramble competition. These results indicate that the pattern of resource acceptance in the presence of conspecifics may be better viewed as a response to increased scramble competition rather than as a response to public information about resource quality.

  16. Helical prospective ECG-gating in cardiac computed tomography: radiation dose and image quality.

    PubMed

    DeFrance, Tony; Dubois, Eric; Gebow, Dan; Ramirez, Alex; Wolf, Florian; Feuchtner, Gudrun M

    2010-01-01

    Helical prospective ECG-gating (pECG) may reduce radiation dose while maintaining the advantages of helical image acquisition for coronary computed tomography angiography (CCTA). Aim of this study was to evaluate helical pECG-gating in CCTA in regards to radiation dose and image quality. 86 patients undergoing 64-multislice CCTA were enrolled. pECG-gating was performed in patients with regular heart rates (HR) < 65 bpm; with the gating window set at 70-85% of the cardiac cycle. All patients received oral and some received additional IV beta-blockers to achieve HR < 65 bpm. In patients with higher or irregular HR, or for functional evaluation, retrospective ECG-gating (rECG) was performed. The average X-ray dose was estimated from the dose length product. Each arterial segment (modified AHA/ACC 17-segment-model) was evaluated on a 4-point image quality scale (4 = excellent; 3 = good, mild artefact; 2 = acceptable, some artefact, 1 = uninterpretable). pECG-gating was applied in 57 patients, rECG-gating in 29 patients. There was no difference in age, gender, body mass index, scan length or tube output settings between both groups. HR in the pECG-group was 54.7 bpm (range, 43-64). The effective radiation dose was significantly lower for patients scanned with pECG-gating with mean 6.9 mSv +/- 1.9 (range, 2.9-10.7) compared to rECG with 16.9 mSv +/- 4.1 (P < 0.001), resulting in a mean dose reduction of 59.2%. For pECG-gating, out of 969 coronary segments, 99.3% were interpretable. Image quality was excellent in 90.2%, good in 7.8%, acceptable in 1.3% and non-interpretable in 0.7% (n = 7 segments). For patients with steady heart rates <65 bpm, helical prospective ECG-gating can significantly lower the radiation dose while maintaining high image quality.

  17. Image quality assessment metric for frame accumulated image

    NASA Astrophysics Data System (ADS)

    Yu, Jianping; Li, Gang; Wang, Shaohui; Lin, Ling

    2018-01-01

    The medical image quality determines the accuracy of diagnosis, and the gray-scale resolution is an important parameter to measure image quality. But current objective metrics are not very suitable for assessing medical images obtained by frame accumulation technology. Little attention was paid to the gray-scale resolution, basically based on spatial resolution and limited to the 256 level gray scale of the existing display device. Thus, this paper proposes a metric, "mean signal-to-noise ratio" (MSNR) based on signal-to-noise in order to be more reasonable to evaluate frame accumulated medical image quality. We demonstrate its potential application through a series of images under a constant illumination signal. Here, the mean image of enough images was regarded as the reference image. Several groups of images by different frame accumulation and their MSNR were calculated. The results of the experiment show that, compared with other quality assessment methods, the metric is simpler, more effective, and more suitable for assessing frame accumulated images that surpass the gray scale and precision of the original image.

  18. Image quality assessment of automatic three-segment MR attenuation correction vs. CT attenuation correction.

    PubMed

    Partovi, Sasan; Kohan, Andres; Gaeta, Chiara; Rubbert, Christian; Vercher-Conejero, Jose L; Jones, Robert S; O'Donnell, James K; Wojtylak, Patrick; Faulhaber, Peter

    2013-01-01

    The purpose of this study is to systematically evaluate the usefulness of Positron emission tomography/Magnetic resonance imaging (PET/MRI) images in a clinical setting by assessing the image quality of Positron emission tomography (PET) images using a three-segment MR attenuation correction (MRAC) versus the standard CT attenuation correction (CTAC). We prospectively studied 48 patients who had their clinically scheduled FDG-PET/CT followed by an FDG-PET/MRI. Three nuclear radiologists evaluated the image quality of CTAC vs. MRAC using a Likert scale (five-point scale). A two-sided, paired t-test was performed for comparison purposes. The image quality was further assessed by categorizing it as acceptable (equal to 4 and 5 on the five-point Likert scale) or unacceptable (equal to 1, 2, and 3 on the five-point Likert scale) quality using the McNemar test. When assessing the image quality using the Likert scale, one reader observed a significant difference between CTAC and MRAC (p=0.0015), whereas the other readers did not observe a difference (p=0.8924 and p=0.1880, respectively). When performing the grouping analysis, no significant difference was found between CTAC vs. MRAC for any of the readers (p=0.6137 for reader 1, p=1 for reader 2, and p=0.8137 for reader 3). All three readers more often reported artifacts on the MRAC images than on the CTAC images. There was no clinically significant difference in quality between PET images generated on a PET/MRI system and those from a Positron emission tomography/Computed tomography (PET/CT) system. PET images using the automatic three-segmented MR attenuation method provided diagnostic image quality. However, future research regarding the image quality obtained using different MR attenuation based methods is warranted before PET/MRI can be used clinically.

  19. Effect of image quality on calcification detection in digital mammography

    PubMed Central

    Warren, Lucy M.; Mackenzie, Alistair; Cooke, Julie; Given-Wilson, Rosalind M.; Wallis, Matthew G.; Chakraborty, Dev P.; Dance, David R.; Bosmans, Hilde; Young, Kenneth C.

    2012-01-01

    (AFROC) area decreased from 0.84 to 0.63 and the ROC area decreased from 0.91 to 0.79 (p < 0.0001). This corresponded to a 30% drop in lesion sensitivity at a NLF equal to 0.1. Detection was also sensitive to the dose used. There was no significant difference in detection between the two image processing algorithms used (p > 0.05). It was additionally found that lower threshold gold thickness from CDMAM analysis implied better cluster detection. The measured threshold gold thickness passed the acceptable limit set in the EU standards for all image qualities except half dose CR. However, calcification detection varied significantly between image qualities. This suggests that the current EU guidelines may need revising. Conclusions: Microcalcification detection was found to be sensitive to detector and dose used. Standard measurements of image quality were a good predictor of microcalcification cluster detection. PMID:22755704

  20. Effect of image quality on calcification detection in digital mammography.

    PubMed

    Warren, Lucy M; Mackenzie, Alistair; Cooke, Julie; Given-Wilson, Rosalind M; Wallis, Matthew G; Chakraborty, Dev P; Dance, David R; Bosmans, Hilde; Young, Kenneth C

    2012-06-01

    0.84 to 0.63 and the ROC area decreased from 0.91 to 0.79 (p < 0.0001). This corresponded to a 30% drop in lesion sensitivity at a NLF equal to 0.1. Detection was also sensitive to the dose used. There was no significant difference in detection between the two image processing algorithms used (p > 0.05). It was additionally found that lower threshold gold thickness from CDMAM analysis implied better cluster detection. The measured threshold gold thickness passed the acceptable limit set in the EU standards for all image qualities except half dose CR. However, calcification detection varied significantly between image qualities. This suggests that the current EU guidelines may need revising. Microcalcification detection was found to be sensitive to detector and dose used. Standard measurements of image quality were a good predictor of microcalcification cluster detection. © 2012 American Association of Physicists in Medicine.

  1. Do Flexible Goal Adjustment and Acceptance Help Preserve Quality of Life in Patients with Multiple Sclerosis?

    PubMed

    Van Damme, Stefaan; De Waegeneer, Annelies; Debruyne, Jan

    2016-06-01

    Goal regulation strategies such as flexible goal adjustment and acceptance are believed to be protective factors in persons with chronic illness, but research on their relative contributions to quality of life in multiple sclerosis (MS) is lacking. We aimed to test the idea that acceptance and flexible goal adjustment (in contrast to tenacious goal pursuit) may help preserve the quality of life in persons with MS. A sample of 117 patients with MS was recruited. They completed questionnaires measuring quality of life (physical functioning, psychological distress), acceptance, flexible goal adjustment, and tenacious goal pursuit. Acceptance significantly accounted for variance in all three indexes of quality of life, beyond the effects of demographic and illness characteristics. The role of goal regulation style was less clear. Flexible goal adjustment significantly accounted for psychological well-being only. Surprisingly, tenacious goal pursuit predicted better psychological functioning and less psychological distress. No support was found for the hypothesis that acceptance and flexible goal adjustment would moderate the relation between illness severity and quality of life. The findings suggest the potential importance of acceptance in understanding MS patients' quality of life, although its hypothesized protective function could not be confirmed. Further conceptual work on acceptance and goal regulation style is needed, as well as prospective work investigating their causal status.

  2. A comparison of defect size and film quality obtained from Film digitized image and digital image radiographs

    NASA Astrophysics Data System (ADS)

    Kamlangkeng, Poramate; Asa, Prateepasen; Mai, Noipitak

    2014-06-01

    Digital radiographic testing is an acceptable premature nondestructive examination technique. Its performance and limitation comparing to the old technique are still not widely well known. In this paper conducted the study on the comparison of the accuracy of the defect size measurement and film quality obtained from film and digital radiograph techniques by testing in specimens and known size sample defect. Initially, one specimen was built with three types of internal defect; which are longitudinal cracking, lack of fusion, and porosity. For the known size sample defect, it was machined various geometrical size for comparing the accuracy of the measuring defect size to the real size in both film and digital images. To compare the image quality by considering at smallest detectable wire and the three defect images. In this research used Image Quality Indicator (IQI) of wire type 10/16 FE EN BS EN-462-1-1994. The radiographic films were produced by X-ray and gamma ray using Kodak AA400 size 3.5x8 inches, while the digital images were produced by Fuji image plate type ST-VI with 100 micrometers resolution. During the tests, a radiator GE model MF3 was implemented. The applied energy is varied from 120 to 220 kV and the current from 1.2 to 3.0 mA. The intensity of Iridium 192 gamma ray is in the range of 24-25 Curie. Under the mentioned conditions, the results showed that the deviation of the defect size measurement comparing to the real size obtained from the digital image radiographs is below than that of the film digitized, whereas the quality of film digitizer radiographs is higher in comparison.

  3. Retinal Image Quality During Accommodation

    PubMed Central

    López-Gil, N.; Martin, J.; Liu, T.; Bradley, A.; Díaz-Muñoz, D.; Thibos, L.

    2013-01-01

    Purpose We asked if retinal image quality is maximum during accommodation, or sub-optimal due to accommodative error, when subjects perform an acuity task. Methods Subjects viewed a monochromatic (552nm), high-contrast letter target placed at various viewing distances. Wavefront aberrations of the accommodating eye were measured near the endpoint of an acuity staircase paradigm. Refractive state, defined as the optimum target vergence for maximising retinal image quality, was computed by through-focus wavefront analysis to find the power of the virtual correcting lens that maximizes visual Strehl ratio. Results Despite changes in ocular aberrations and pupil size during binocular viewing, retinal image quality and visual acuity typically remain high for all target vergences. When accommodative errors lead to sub-optimal retinal image quality, acuity and measured image quality both decline. However, the effect of accommodation errors of on visual acuity are mitigated by pupillary constriction associated with accommodation and binocular convergence and also to binocular summation of dissimilar retinal image blur. Under monocular viewing conditions some subjects displayed significant accommodative lag that reduced visual performance, an effect that was exacerbated by pharmacological dilation of the pupil. Conclusions Spurious measurement of accommodative error can be avoided when the image quality metric used to determine refractive state is compatible with the focusing criteria used by the visual system to control accommodation. Real focusing errors of the accommodating eye do not necessarily produce a reliably measurable loss of image quality or clinically significant loss of visual performance, probably because of increased depth-of-focus due to pupil constriction. When retinal image quality is close to maximum achievable (given the eye’s higher-order aberrations), acuity is also near maximum. A combination of accommodative lag, reduced image quality, and reduced

  4. JPEG2000 still image coding quality.

    PubMed

    Chen, Tzong-Jer; Lin, Sheng-Chieh; Lin, You-Chen; Cheng, Ren-Gui; Lin, Li-Hui; Wu, Wei

    2013-10-01

    This work demonstrates the image qualities between two popular JPEG2000 programs. Two medical image compression algorithms are both coded using JPEG2000, but they are different regarding the interface, convenience, speed of computation, and their characteristic options influenced by the encoder, quantization, tiling, etc. The differences in image quality and compression ratio are also affected by the modality and compression algorithm implementation. Do they provide the same quality? The qualities of compressed medical images from two image compression programs named Apollo and JJ2000 were evaluated extensively using objective metrics. These algorithms were applied to three medical image modalities at various compression ratios ranging from 10:1 to 100:1. Following that, the quality of the reconstructed images was evaluated using five objective metrics. The Spearman rank correlation coefficients were measured under every metric in the two programs. We found that JJ2000 and Apollo exhibited indistinguishable image quality for all images evaluated using the above five metrics (r > 0.98, p < 0.001). It can be concluded that the image quality of the JJ2000 and Apollo algorithms is statistically equivalent for medical image compression.

  5. Process perspective on image quality evaluation

    NASA Astrophysics Data System (ADS)

    Leisti, Tuomas; Halonen, Raisa; Kokkonen, Anna; Weckman, Hanna; Mettänen, Marja; Lensu, Lasse; Ritala, Risto; Oittinen, Pirkko; Nyman, Göte

    2008-01-01

    The psychological complexity of multivariate image quality evaluation makes it difficult to develop general image quality metrics. Quality evaluation includes several mental processes and ignoring these processes and the use of a few test images can lead to biased results. By using a qualitative/quantitative (Interpretation Based Quality, IBQ) methodology, we examined the process of pair-wise comparison in a setting, where the quality of the images printed by laser printer on different paper grades was evaluated. Test image consisted of a picture of a table covered with several objects. Three other images were also used, photographs of a woman, cityscape and countryside. In addition to the pair-wise comparisons, observers (N=10) were interviewed about the subjective quality attributes they used in making their quality decisions. An examination of the individual pair-wise comparisons revealed serious inconsistencies in observers' evaluations on the test image content, but not on other contexts. The qualitative analysis showed that this inconsistency was due to the observers' focus of attention. The lack of easily recognizable context in the test image may have contributed to this inconsistency. To obtain reliable knowledge of the effect of image context or attention on subjective image quality, a qualitative methodology is needed.

  6. Retinal image quality during accommodation.

    PubMed

    López-Gil, Norberto; Martin, Jesson; Liu, Tao; Bradley, Arthur; Díaz-Muñoz, David; Thibos, Larry N

    2013-07-01

    We asked if retinal image quality is maximum during accommodation, or sub-optimal due to accommodative error, when subjects perform an acuity task. Subjects viewed a monochromatic (552 nm), high-contrast letter target placed at various viewing distances. Wavefront aberrations of the accommodating eye were measured near the endpoint of an acuity staircase paradigm. Refractive state, defined as the optimum target vergence for maximising retinal image quality, was computed by through-focus wavefront analysis to find the power of the virtual correcting lens that maximizes visual Strehl ratio. Despite changes in ocular aberrations and pupil size during binocular viewing, retinal image quality and visual acuity typically remain high for all target vergences. When accommodative errors lead to sub-optimal retinal image quality, acuity and measured image quality both decline. However, the effect of accommodation errors of on visual acuity are mitigated by pupillary constriction associated with accommodation and binocular convergence and also to binocular summation of dissimilar retinal image blur. Under monocular viewing conditions some subjects displayed significant accommodative lag that reduced visual performance, an effect that was exacerbated by pharmacological dilation of the pupil. Spurious measurement of accommodative error can be avoided when the image quality metric used to determine refractive state is compatible with the focusing criteria used by the visual system to control accommodation. Real focusing errors of the accommodating eye do not necessarily produce a reliably measurable loss of image quality or clinically significant loss of visual performance, probably because of increased depth-of-focus due to pupil constriction. When retinal image quality is close to maximum achievable (given the eye's higher-order aberrations), acuity is also near maximum. A combination of accommodative lag, reduced image quality, and reduced visual function may be a useful

  7. Image quality scaling of electrophotographic prints

    NASA Astrophysics Data System (ADS)

    Johnson, Garrett M.; Patil, Rohit A.; Montag, Ethan D.; Fairchild, Mark D.

    2003-12-01

    Two psychophysical experiments were performed scaling overall image quality of black-and-white electrophotographic (EP) images. Six different printers were used to generate the images. There were six different scenes included in the experiment, representing photographs, business graphics, and test-targets. The two experiments were split into a paired-comparison experiment examining overall image quality, and a triad experiment judging overall similarity and dissimilarity of the printed images. The paired-comparison experiment was analyzed using Thurstone's Law, to generate an interval scale of quality, and with dual scaling, to determine the independent dimensions used for categorical scaling. The triad experiment was analyzed using multidimensional scaling to generate a psychological stimulus space. The psychophysical results indicated that the image quality was judged mainly along one dimension and that the relationships among the images can be described with a single dimension in most cases. Regression of various physical measurements of the images to the paired comparison results showed that a small number of physical attributes of the images could be correlated with the psychophysical scale of image quality. However, global image difference metrics did not correlate well with image quality.

  8. SU-F-I-47: Optimizing Protocols for Image Quality and Dose in Abdominal CT of Large Patients

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Johnson, L; Yester, M

    Purpose: Newer CT scanners are able to use scout views to adjust mA throughout the scan in order to achieve a given noise level. However, given constraints of radiologist preferences for kVp and rotation time, it may not be possible to achieve an acceptable noise level for large patients. A study was initiated to determine for which patients kVp and/or rotation time should be changed in order to achieve acceptable image quality. Methods: Patient scans were reviewed on two new Emergency Department scanners (Philips iCT) to identify patients over a large range of sizes. These iCTs were set with amore » limit of 500 mA to safeguard against a failure that might cause a CT scan to be (incorrectly) obtained at too-high mA. Scout views of these scans were assessed for both AP and LAT patient width and AP and LAT standard deviation in an ROI over the liver. Effective diameter and product of the scout standard deviations over the liver were both studied as possible metrics for identifying patients who would need kVp and/or rotation time changed. The mA used for the liver in the CT was compared to these metrics for those patients whose CT scans showed acceptable image quality. Results: Both effective diameter and product of the scout standard deviations over the liver result in similar predictions for which patients will require the kVp and/or rotation time to be changed to achieve an optimal combination of image quality and dose. Conclusion: Two mechanisms for CT technologists to determine based on scout characteristics what kVp, mA limit, and rotation time to use when DoseRight with our physicians’ preferred kVp and rotation time will not yield adequate image quality are described.« less

  9. A comprehensive study on the relationship between the image quality and imaging dose in low-dose cone beam CT

    NASA Astrophysics Data System (ADS)

    Yan, Hao; Cervino, Laura; Jia, Xun; Jiang, Steve B.

    2012-04-01

    While compressed sensing (CS)-based algorithms have been developed for the low-dose cone beam CT (CBCT) reconstruction, a clear understanding of the relationship between the image quality and imaging dose at low-dose levels is needed. In this paper, we qualitatively investigate this subject in a comprehensive manner with extensive experimental and simulation studies. The basic idea is to plot both the image quality and imaging dose together as functions of the number of projections and mAs per projection over the whole clinically relevant range. On this basis, a clear understanding of the tradeoff between the image quality and imaging dose can be achieved and optimal low-dose CBCT scan protocols can be developed to maximize the dose reduction while minimizing the image quality loss for various imaging tasks in image-guided radiation therapy (IGRT). Main findings of this work include (1) under the CS-based reconstruction framework, image quality has little degradation over a large range of dose variation. Image quality degradation becomes evident when the imaging dose (approximated with the x-ray tube load) is decreased below 100 total mAs. An imaging dose lower than 40 total mAs leads to a dramatic image degradation, and thus should be used cautiously. Optimal low-dose CBCT scan protocols likely fall in the dose range of 40-100 total mAs, depending on the specific IGRT applications. (2) Among different scan protocols at a constant low-dose level, the super sparse-view reconstruction with the projection number less than 50 is the most challenging case, even with strong regularization. Better image quality can be acquired with low mAs protocols. (3) The optimal scan protocol is the combination of a medium number of projections and a medium level of mAs/view. This is more evident when the dose is around 72.8 total mAs or below and when the ROI is a low-contrast or high-resolution object. Based on our results, the optimal number of projections is around 90 to 120. (4

  10. MR colonography with fecal tagging: do individual patient characteristics influence image quality?

    PubMed

    Kinner, Sonja; Kuehle, Christiane A; Langhorst, Jost; Ladd, Susanne C; Nuefer, Michael; Barkhausen, Joerg; Lauenstein, Thomas C

    2007-05-01

    To evaluate if different patient characteristics influence performance of fecal tagging (a new MR colonography (MRC) technique to label stool to avoid bowel cleansing) and, consecutively, MR image quality. A total of 333 patients (mean age = 61 years) underwent MRC with fecal tagging. Four segments of the large bowel (ascending, transverse, descending, and sigmoid colon) were assessed as for the presence of nontagged stool particles, which can impede an assessment of the colonic wall. Ratings were correlated with patients' characteristics including patient age (<55 vs. > or =55 years), body mass index (BMI) (<25 vs. > or =25), gender, and acceptance levels for fecal tagging. Statistical analysis was performed using a Mann-Whitney U-test. A total of 1332 colonic segments were evaluated. Among them, 327 segments (25%) did not contain any visible stool particles. Considerably reduced image quality was found in 61 segments (5%). Best image quality was found in the sigmoid colon (mean value = 1.9), while image quality of the ascending colon turned out to be worst (mean value = 2.6). Fecal tagging effectiveness showed a reverse correlation with patient age. However, all other characteristics did not have a statistically significant influence on fecal tagging outcome. MRC in conjunction with barium-based fecal tagging led to diagnostic image quality in 95% of all colonic segments. Since tagging results were significantly decreased in patients > or =55 years, tagging protocols should to be modified in this group, i.e., by increasing the time interval of tagging administration. (c) 2007 Wiley-Liss, Inc.

  11. Influence of physical parameters on radiation protection and image quality in intra-oral radiology

    NASA Astrophysics Data System (ADS)

    Belinato, W.; Souza, D. N.

    2011-10-01

    In the world of diagnostic imaging, radiography is an important supplementary method for dental diagnosis. In radiology, special attention must be paid to the radiological protection of patients and health professionals, and also to image quality for correct diagnosis. In Brazil, the national rules governing the operation of medical and dental radiology were specified in 1998 by the National Sanitary Surveillance Agency, complemented in 2005 by the guide "Medical radiology: security and performance of equipment." In this study, quality control tests were performed in public clinics with dental X-ray equipment in the State of Sergipe, Brazil, with consideration of the physical parameters that influence radiological protection and also the quality of images taken in intra-oral radiography. The accuracy of the exposure time was considered acceptable for equipment with digital timers. Exposure times and focal-spot size variations can lead to increased entrance dose. Increased dose has also been associated with visual processing of radiographic film, which often requires repeating the radiographic examination.

  12. Process Improvements in Training Device Acceptance Testing: A Study in Total Quality Management

    DTIC Science & Technology

    1990-12-12

    Quality Management , a small group of Government and industry specialists examined the existing training device acceptance test process for potential improvements. The agreed-to mission of the Air Force/Industry partnership was to continuously identify and promote implementable approaches to minimize the cost and time required for acceptance testing while ensuring that validated performance supports the user training requirements. Application of a Total Quality process improvement model focused on the customers and their requirements, analyzed how work was accomplished, and

  13. Limiting Magnitude, τ, t eff, and Image Quality in DES Year 1

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    H. Neilsen, Jr.; Bernstein, Gary; Gruendl, Robert

    The Dark Energy Survey (DES) is an astronomical imaging survey being completed with the DECam imager on the Blanco telescope at CTIO. After each night of observing, the DES data management (DM) group performs an initial processing of that night's data, and uses the results to determine which exposures are of acceptable quality, and which need to be repeated. The primary measure by which we declare an image of acceptable quality ismore » $$\\tau$$, a scaling of the exposure time. This is the scale factor that needs to be applied to the open shutter time to reach the same photometric signal to noise ratio for faint point sources under a set of canonical good conditions. These conditions are defined to be seeing resulting in a PSF full width at half maximum (FWHM) of 0.9" and a pre-defined sky brightness which approximates the zenith sky brightness under fully dark conditions. Point source limiting magnitude and signal to noise should therefore vary with t in the same way they vary with exposure time. Measurements of point sources and $$\\tau$$ in the first year of DES data confirm that they do. In the context of DES, the symbol $$t_{eff}$$ and the expression "effective exposure time" usually refer to the scaling factor, $$\\tau$$, rather than the actual effective exposure time; the "effective exposure time" in this case refers to the effective duration of one second, rather than the effective duration of an exposure.« less

  14. Quality Control of Structural MRI Images Applied Using FreeSurfer—A Hands-On Workflow to Rate Motion Artifacts

    PubMed Central

    Backhausen, Lea L.; Herting, Megan M.; Buse, Judith; Roessner, Veit; Smolka, Michael N.; Vetter, Nora C.

    2016-01-01

    In structural magnetic resonance imaging motion artifacts are common, especially when not scanning healthy young adults. It has been shown that motion affects the analysis with automated image-processing techniques (e.g., FreeSurfer). This can bias results. Several developmental and adult studies have found reduced volume and thickness of gray matter due to motion artifacts. Thus, quality control is necessary in order to ensure an acceptable level of quality and to define exclusion criteria of images (i.e., determine participants with most severe artifacts). However, information about the quality control workflow and image exclusion procedure is largely lacking in the current literature and the existing rating systems differ. Here, we propose a stringent workflow of quality control steps during and after acquisition of T1-weighted images, which enables researchers dealing with populations that are typically affected by motion artifacts to enhance data quality and maximize sample sizes. As an underlying aim we established a thorough quality control rating system for T1-weighted images and applied it to the analysis of developmental clinical data using the automated processing pipeline FreeSurfer. This hands-on workflow and quality control rating system will aid researchers in minimizing motion artifacts in the final data set, and therefore enhance the quality of structural magnetic resonance imaging studies. PMID:27999528

  15. Accounting for patient size in the optimization of dose and image quality of pelvis cone beam CT protocols on the Varian OBI system

    PubMed Central

    Moore, Craig S; Horsfield, Carl J; Saunderson, John R; Beavis, Andrew W

    2015-01-01

    Objective: The purpose of this study was to develop size-based radiotherapy kilovoltage cone beam CT (CBCT) protocols for the pelvis. Methods: Image noise was measured in an elliptical phantom of varying size for a range of exposure factors. Based on a previously defined “small pelvis” reference patient and CBCT protocol, appropriate exposure factors for small, medium, large and extra-large patients were derived which approximate the image noise behaviour observed on a Philips CT scanner (Philips Medical Systems, Best, Netherlands) with automatic exposure control (AEC). Selection criteria, based on maximum tube current–time product per rotation selected during the radiotherapy treatment planning scan, were derived based on an audit of patient size. Results: It has been demonstrated that 110 kVp yields acceptable image noise for reduced patient dose in pelvic CBCT scans of small, medium and large patients, when compared with manufacturer's default settings (125 kVp). Conversely, extra-large patients require increased exposure factors to give acceptable images. 57% of patients in the local population now receive much lower radiation doses, whereas 13% require higher doses (but now yield acceptable images). Conclusion: The implementation of size-based exposure protocols has significantly reduced radiation dose to the majority of patients with no negative impact on image quality. Increased doses are required on the largest patients to give adequate image quality. Advances in knowledge: The development of size-based CBCT protocols that use the planning CT scan (with AEC) to determine which protocol is appropriate ensures adequate image quality whilst minimizing patient radiation dose. PMID:26419892

  16. Accounting for patient size in the optimization of dose and image quality of pelvis cone beam CT protocols on the Varian OBI system.

    PubMed

    Wood, Tim J; Moore, Craig S; Horsfield, Carl J; Saunderson, John R; Beavis, Andrew W

    2015-01-01

    The purpose of this study was to develop size-based radiotherapy kilovoltage cone beam CT (CBCT) protocols for the pelvis. Image noise was measured in an elliptical phantom of varying size for a range of exposure factors. Based on a previously defined "small pelvis" reference patient and CBCT protocol, appropriate exposure factors for small, medium, large and extra-large patients were derived which approximate the image noise behaviour observed on a Philips CT scanner (Philips Medical Systems, Best, Netherlands) with automatic exposure control (AEC). Selection criteria, based on maximum tube current-time product per rotation selected during the radiotherapy treatment planning scan, were derived based on an audit of patient size. It has been demonstrated that 110 kVp yields acceptable image noise for reduced patient dose in pelvic CBCT scans of small, medium and large patients, when compared with manufacturer's default settings (125 kVp). Conversely, extra-large patients require increased exposure factors to give acceptable images. 57% of patients in the local population now receive much lower radiation doses, whereas 13% require higher doses (but now yield acceptable images). The implementation of size-based exposure protocols has significantly reduced radiation dose to the majority of patients with no negative impact on image quality. Increased doses are required on the largest patients to give adequate image quality. The development of size-based CBCT protocols that use the planning CT scan (with AEC) to determine which protocol is appropriate ensures adequate image quality whilst minimizing patient radiation dose.

  17. Assessing product image quality for online shopping

    NASA Astrophysics Data System (ADS)

    Goswami, Anjan; Chung, Sung H.; Chittar, Naren; Islam, Atiq

    2012-01-01

    Assessing product-image quality is important in the context of online shopping. A high quality image that conveys more information about a product can boost the buyer's confidence and can get more attention. However, the notion of image quality for product-images is not the same as that in other domains. The perception of quality of product-images depends not only on various photographic quality features but also on various high level features such as clarity of the foreground or goodness of the background etc. In this paper, we define a notion of product-image quality based on various such features. We conduct a crowd-sourced experiment to collect user judgments on thousands of eBay's images. We formulate a multi-class classification problem for modeling image quality by classifying images into good, fair and poor quality based on the guided perceptual notions from the judges. We also conduct experiments with regression using average crowd-sourced human judgments as target. We compute a pseudo-regression score with expected average of predicted classes and also compute a score from the regression technique. We design many experiments with various sampling and voting schemes with crowd-sourced data and construct various experimental image quality models. Most of our models have reasonable accuracies (greater or equal to 70%) on test data set. We observe that our computed image quality score has a high (0.66) rank correlation with average votes from the crowd sourced human judgments.

  18. SU-E-J-06: A Time Dependence Analysis of CBCT Image Quality and Mechanical Stability.

    PubMed

    Oves, S; Stenbeck, J; Gebreamlak, W; Alkhatib, H

    2012-06-01

    To quantify the change, if any, in flexmap correction factors and image quality with the XVI system over a course of several years and from these results, assess their clinical impact. Flexmap, a calibration procedure which corrects for imperfect gantry rotation for cone-beam CT reconstruction, and image quality tests were performed on three Elekta Synergy linacs equipped with XVI. Data was collected per month over three years. U and V values, corresponding to lateral and longitudinal shifts respectively, were acquired through the XVI software. Image quality parameters were obtained through CT imaging of the Catphan 500®. For each reconstruction, pixel values for low density polyethylene (LDPE) and polystyrene materials were recorded. For all three linacs, analysis of the flexmap showed a significant change in the U factor for both month-to-month comparisons and comparisons between machines. The V correction factor exhibited a small variation month to month, and showed a slight, gradual increase over time (0.2 +/-0.08 mm). Image quality analysis showed a near consistent decrease (5-10%) in LDPE and polystyrene. Despite this decrease in pixel values, the ratio of the two pixel values remained constant, thus a similar decreasing trend in contrast was not observed. Analysis of monthly flexmap calibration showed the general monthly change in correction shifts and their general trend over several years. For image quality, our research exhibited roughly 0.5% per month decrease in pixel values of the Catphan®. Our results imply that CBCT images obtained from XVI are not appropriate for treatment planning and despite the decrease in panel response over time, image quality with respect to contrast will remain within acceptable clinical standards. Future studies may be carried out to assess any correlation between image quality and XVI source strength. © 2012 American Association of Physicists in Medicine.

  19. A qualitative and quantitative analysis of radiation dose and image quality of computed tomography images using adaptive statistical iterative reconstruction.

    PubMed

    Hussain, Fahad Ahmed; Mail, Noor; Shamy, Abdulrahman M; Suliman, Alghamdi; Saoudi, Abdelhamid

    2016-05-08

    Image quality is a key issue in radiology, particularly in a clinical setting where it is important to achieve accurate diagnoses while minimizing radiation dose. Some computed tomography (CT) manufacturers have introduced algorithms that claim significant dose reduction. In this study, we assessed CT image quality produced by two reconstruction algorithms provided with GE Healthcare's Discovery 690 Elite positron emission tomography (PET) CT scanner. Image quality was measured for images obtained at various doses with both conventional filtered back-projection (FBP) and adaptive statistical iterative reconstruction (ASIR) algorithms. A stan-dard CT dose index (CTDI) phantom and a pencil ionization chamber were used to measure the CT dose at 120 kVp and an exposure of 260 mAs. Image quality was assessed using two phantoms. CT images of both phantoms were acquired at tube voltage (kV) of 120 with exposures ranging from 25 mAs to 400 mAs. Images were reconstructed using FBP and ASIR ranging from 10% to 100%, then analyzed for noise, low-contrast detectability, contrast-to-noise ratio (CNR), and modulation transfer function (MTF). Noise was 4.6 HU in water phantom images acquired at 260 mAs/FBP 120 kV and 130 mAs/50% ASIR 120 kV. The large objects (fre-quency < 7 lp/cm) retained fairly acceptable image quality at 130 mAs/50% ASIR, compared to 260 mAs/FBP. The application of ASIR for small objects (frequency >7 lp/cm) showed poor visibility compared to FBP at 260 mAs and even worse for images acquired at less than 130 mAs. ASIR blending more than 50% at low dose tends to reduce contrast of small objects (frequency >7 lp/cm). We concluded that dose reduction and ASIR should be applied with close attention if the objects to be detected or diagnosed are small (frequency > 7 lp/cm). Further investigations are required to correlate the small objects (frequency > 7 lp/cm) to patient anatomy and clinical diagnosis.

  20. Image Quality Ranking Method for Microscopy

    PubMed Central

    Koho, Sami; Fazeli, Elnaz; Eriksson, John E.; Hänninen, Pekka E.

    2016-01-01

    Automated analysis of microscope images is necessitated by the increased need for high-resolution follow up of events in time. Manually finding the right images to be analyzed, or eliminated from data analysis are common day-to-day problems in microscopy research today, and the constantly growing size of image datasets does not help the matter. We propose a simple method and a software tool for sorting images within a dataset, according to their relative quality. We demonstrate the applicability of our method in finding good quality images in a STED microscope sample preparation optimization image dataset. The results are validated by comparisons to subjective opinion scores, as well as five state-of-the-art blind image quality assessment methods. We also show how our method can be applied to eliminate useless out-of-focus images in a High-Content-Screening experiment. We further evaluate the ability of our image quality ranking method to detect out-of-focus images, by extensive simulations, and by comparing its performance against previously published, well-established microscopy autofocus metrics. PMID:27364703

  1. 40 CFR 90.510 - Compliance with acceptable quality level and passing and failing criteria for selective...

    Code of Federal Regulations, 2011 CFR

    2011-07-01

    ... 40 Protection of Environment 20 2011-07-01 2011-07-01 false Compliance with acceptable quality level and passing and failing criteria for selective enforcement audits. 90.510 Section 90.510....510 Compliance with acceptable quality level and passing and failing criteria for selective...

  2. Comparative Analysis of Reconstructed Image Quality in a Simulated Chromotomographic Imager

    DTIC Science & Technology

    2014-03-01

    quality . This example uses five basic images a backlit bar chart with random intensity, 100 nm separation. A total of 54 initial target...compared for a variety of scenes. Reconstructed image quality is highly dependent on the initial target hypercube so a total of 54 initial target...COMPARATIVE ANALYSIS OF RECONSTRUCTED IMAGE QUALITY IN A SIMULATED CHROMOTOMOGRAPHIC IMAGER THESIS

  3. Global quality imaging: emerging issues.

    PubMed

    Lau, Lawrence S; Pérez, Maria R; Applegate, Kimberly E; Rehani, Madan M; Ringertz, Hans G; George, Robert

    2011-07-01

    Quality imaging may be described as "a timely access to and delivery of integrated and appropriate procedures, in a safe and responsive practice, and a prompt delivery of an accurately interpreted report by capable personnel in an efficient, effective, and sustainable manner." For this article, radiation safety is considered as one of the key quality elements. The stakeholders are the drivers of quality imaging. These include those that directly provide or use imaging procedures and others indirectly supporting the system. Imaging is indispensable in health care, and its use has greatly expanded worldwide. Globalization, consumer sophistication, communication and technological advances, corporatization, rationalization, service outsourcing, teleradiology, workflow modularization, and commoditization are reshaping practice. This article defines the emerging issues; an earlier article in the May 2011 issue described possible improvement actions. The issues that could threaten the quality use of imaging for all countries include workforce shortage; increased utilization, population radiation exposure, and cost; practice changes; and efficiency drive and budget constraints. In response to these issues, a range of quality improvement measures, strategies, and actions are used to maximize the benefits and minimize the risks. The 3 measures are procedure justification, optimization of image quality and radiation protection, and error prevention. The development and successful implementation of such improvement actions require leadership, collaboration, and the active participation of all stakeholders to achieve the best outcomes that we all advocate. Copyright © 2011 American College of Radiology. Published by Elsevier Inc. All rights reserved.

  4. Automatic quality assessment of planetary images

    NASA Astrophysics Data System (ADS)

    Sidiropoulos, P.; Muller, J.-P.

    2015-10-01

    A significant fraction of planetary images are corrupted beyond the point that much scientific meaning can be extracted. For example, transmission errors result in missing data which is unrecoverable. The available planetary image datasets include many such "bad data", which both occupy valuable scientific storage resources and create false impressions about planetary image availability for specific planetary objects or target areas. In this work, we demonstrate a pipeline that we have developed to automatically assess the quality of planetary images. Additionally, this method discriminates between different types of image degradation, such as low-quality originating from camera flaws or low-quality triggered by atmospheric conditions, etc. Examples of quality assessment results for Viking Orbiter imagery will be also presented.

  5. Guest Editorial Image Quality

    NASA Astrophysics Data System (ADS)

    Cheatham, Patrick S.

    1982-02-01

    The term image quality can, unfortunately, apply to anything from a public relations firm's discussion to a comparison between corner drugstores' film processing. If we narrow the discussion to optical systems, we clarify the problem somewhat, but only slightly. We are still faced with a multitude of image quality measures all different, and all couched in different terminology. Optical designers speak of MTF values, digital processors talk about summations of before and after image differences, pattern recognition engineers allude to correlation values, and radar imagers use side-lobe response values measured in decibels. Further complexity is introduced by terms such as information content, bandwidth, Strehl ratios, and, of course, limiting resolution. The problem is to compare these different yardsticks and try to establish some concrete ideas about evaluation of a final image. We need to establish the image attributes which are the most important to perception of the image in question and then begin to apply the different system parameters to those attributes.

  6. 40 CFR 91.608 - Compliance with acceptable quality level and passing and failing criteria for selective...

    Code of Federal Regulations, 2011 CFR

    2011-07-01

    ... level and passing and failing criteria for selective enforcement audits. 91.608 Section 91.608... with acceptable quality level and passing and failing criteria for selective enforcement audits. (a) The prescribed acceptable quality level is 40 percent. (b) A failed engine is one whose final test...

  7. 40 CFR 89.510 - Compliance with acceptable quality level and passing and failing criteria for selective...

    Code of Federal Regulations, 2013 CFR

    2013-07-01

    ... level and passing and failing criteria for selective enforcement audits. 89.510 Section 89.510... Compliance with acceptable quality level and passing and failing criteria for selective enforcement audits. (a) The prescribed acceptable quality level is 40 percent. (b) A failed engine is one whose final...

  8. 40 CFR 89.510 - Compliance with acceptable quality level and passing and failing criteria for selective...

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... level and passing and failing criteria for selective enforcement audits. 89.510 Section 89.510... Compliance with acceptable quality level and passing and failing criteria for selective enforcement audits. (a) The prescribed acceptable quality level is 40 percent. (b) A failed engine is one whose final...

  9. 40 CFR 91.608 - Compliance with acceptable quality level and passing and failing criteria for selective...

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... level and passing and failing criteria for selective enforcement audits. 91.608 Section 91.608... with acceptable quality level and passing and failing criteria for selective enforcement audits. (a) The prescribed acceptable quality level is 40 percent. (b) A failed engine is one whose final test...

  10. 40 CFR 91.608 - Compliance with acceptable quality level and passing and failing criteria for selective...

    Code of Federal Regulations, 2013 CFR

    2013-07-01

    ... level and passing and failing criteria for selective enforcement audits. 91.608 Section 91.608... with acceptable quality level and passing and failing criteria for selective enforcement audits. (a) The prescribed acceptable quality level is 40 percent. (b) A failed engine is one whose final test...

  11. Online Support Service Quality, Online Learning Acceptance, and Student Satisfaction

    ERIC Educational Resources Information Center

    Lee, Jung-Wan

    2010-01-01

    This paper examines potential differences between Korean and American students in terms of their perception levels regarding online education support service quality, online learning acceptance, and satisfaction. Eight hundred and seventy-two samples, which were collected from students in online classes in the United States and Korea, were…

  12. An Underwater Color Image Quality Evaluation Metric.

    PubMed

    Yang, Miao; Sowmya, Arcot

    2015-12-01

    Quality evaluation of underwater images is a key goal of underwater video image retrieval and intelligent processing. To date, no metric has been proposed for underwater color image quality evaluation (UCIQE). The special absorption and scattering characteristics of the water medium do not allow direct application of natural color image quality metrics especially to different underwater environments. In this paper, subjective testing for underwater image quality has been organized. The statistical distribution of the underwater image pixels in the CIELab color space related to subjective evaluation indicates the sharpness and colorful factors correlate well with subjective image quality perception. Based on these, a new UCIQE metric, which is a linear combination of chroma, saturation, and contrast, is proposed to quantify the non-uniform color cast, blurring, and low-contrast that characterize underwater engineering and monitoring images. Experiments are conducted to illustrate the performance of the proposed UCIQE metric and its capability to measure the underwater image enhancement results. They show that the proposed metric has comparable performance to the leading natural color image quality metrics and the underwater grayscale image quality metrics available in the literature, and can predict with higher accuracy the relative amount of degradation with similar image content in underwater environments. Importantly, UCIQE is a simple and fast solution for real-time underwater video processing. The effectiveness of the presented measure is also demonstrated by subjective evaluation. The results show better correlation between the UCIQE and the subjective mean opinion score.

  13. Blind image quality assessment based on aesthetic and statistical quality-aware features

    NASA Astrophysics Data System (ADS)

    Jenadeleh, Mohsen; Masaeli, Mohammad Masood; Moghaddam, Mohsen Ebrahimi

    2017-07-01

    The main goal of image quality assessment (IQA) methods is the emulation of human perceptual image quality judgments. Therefore, the correlation between objective scores of these methods with human perceptual scores is considered as their performance metric. Human judgment of the image quality implicitly includes many factors when assessing perceptual image qualities such as aesthetics, semantics, context, and various types of visual distortions. The main idea of this paper is to use a host of features that are commonly employed in image aesthetics assessment in order to improve blind image quality assessment (BIQA) methods accuracy. We propose an approach that enriches the features of BIQA methods by integrating a host of aesthetics image features with the features of natural image statistics derived from multiple domains. The proposed features have been used for augmenting five different state-of-the-art BIQA methods, which use statistical natural scene statistics features. Experiments were performed on seven benchmark image quality databases. The experimental results showed significant improvement of the accuracy of the methods.

  14. Image quality and dose differences caused by vendor-specific image processing of neonatal radiographs.

    PubMed

    Sensakovic, William F; O'Dell, M Cody; Letter, Haley; Kohler, Nathan; Rop, Baiywo; Cook, Jane; Logsdon, Gregory; Varich, Laura

    2016-10-01

    Image processing plays an important role in optimizing image quality and radiation dose in projection radiography. Unfortunately commercial algorithms are black boxes that are often left at or near vendor default settings rather than being optimized. We hypothesize that different commercial image-processing systems, when left at or near default settings, create significant differences in image quality. We further hypothesize that image-quality differences can be exploited to produce images of equivalent quality but lower radiation dose. We used a portable radiography system to acquire images on a neonatal chest phantom and recorded the entrance surface air kerma (ESAK). We applied two image-processing systems (Optima XR220amx, by GE Healthcare, Waukesha, WI; and MUSICA(2) by Agfa HealthCare, Mortsel, Belgium) to the images. Seven observers (attending pediatric radiologists and radiology residents) independently assessed image quality using two methods: rating and matching. Image-quality ratings were independently assessed by each observer on a 10-point scale. Matching consisted of each observer matching GE-processed images and Agfa-processed images with equivalent image quality. A total of 210 rating tasks and 42 matching tasks were performed and effective dose was estimated. Median Agfa-processed image-quality ratings were higher than GE-processed ratings. Non-diagnostic ratings were seen over a wider range of doses for GE-processed images than for Agfa-processed images. During matching tasks, observers matched image quality between GE-processed images and Agfa-processed images acquired at a lower effective dose (11 ± 9 μSv; P < 0.0001). Image-processing methods significantly impact perceived image quality. These image-quality differences can be exploited to alter protocols and produce images of equivalent image quality but lower doses. Those purchasing projection radiography systems or third-party image-processing software should be aware that image

  15. The acceptance of illness, the intensity of pain and the quality of life in patients with lung cancer

    PubMed Central

    Polański, Jacek; Jankowska-Polanska, Beata; Lomper, Katarzyna; Janczak, Dariusz; Rosinczuk, Joanna

    2017-01-01

    Background Lung cancer is the major cause of cancer related deaths worldwide. The overall 5-year survival rate is very low and accounts for only 15%. Poor quality of life is considered a prognostic factor for shorter survival in lung cancer patients. The aim of the study was to examine the relationships between pain, the acceptance of illness and quality of life in patients with lung cancer. Methods The study included 155 patients with lung cancer with mean age of 62.23 [standard deviation (SD)=9.86] years. We used the Acceptance of Illness Scale (AIS) and the Visual Analog Scale (VAS) for pain, and the Short Form Health Survey (SF-8) for the assessment of quality of life. For statistical analysis, Spearman’s rank correlation coefficient and linear regression method were used. Results Mean score of the acceptance of illness was 27.1 (SD=9.2). Mean score of the pain severity as measured by the VAS was 4.07 (SD=1.83). The acceptance of illness was significantly positively correlated with all the domains of quality of life. Both the AIS and the VAS were independent determinants of physical and mental components of quality of life. Age and World Health Organization (WHO) performance status were additional predictors of physical component of quality of life. Conclusions The knowledge about the acceptance of illness, consequences of the decreased acceptance of illness, and factors affecting its level in patients with lung cancer is still insufficient. Relationships among acceptance of illness, quality of life, and pain should be further investigated. PMID:29221267

  16. The acceptance of illness, the intensity of pain and the quality of life in patients with lung cancer.

    PubMed

    Chabowski, Mariusz; Polański, Jacek; Jankowska-Polanska, Beata; Lomper, Katarzyna; Janczak, Dariusz; Rosinczuk, Joanna

    2017-09-01

    Lung cancer is the major cause of cancer related deaths worldwide. The overall 5-year survival rate is very low and accounts for only 15%. Poor quality of life is considered a prognostic factor for shorter survival in lung cancer patients. The aim of the study was to examine the relationships between pain, the acceptance of illness and quality of life in patients with lung cancer. The study included 155 patients with lung cancer with mean age of 62.23 [standard deviation (SD)=9.86] years. We used the Acceptance of Illness Scale (AIS) and the Visual Analog Scale (VAS) for pain, and the Short Form Health Survey (SF-8) for the assessment of quality of life. For statistical analysis, Spearman's rank correlation coefficient and linear regression method were used. Mean score of the acceptance of illness was 27.1 (SD=9.2). Mean score of the pain severity as measured by the VAS was 4.07 (SD=1.83). The acceptance of illness was significantly positively correlated with all the domains of quality of life. Both the AIS and the VAS were independent determinants of physical and mental components of quality of life. Age and World Health Organization (WHO) performance status were additional predictors of physical component of quality of life. The knowledge about the acceptance of illness, consequences of the decreased acceptance of illness, and factors affecting its level in patients with lung cancer is still insufficient. Relationships among acceptance of illness, quality of life, and pain should be further investigated.

  17. Image quality assessment using deep convolutional networks

    NASA Astrophysics Data System (ADS)

    Li, Yezhou; Ye, Xiang; Li, Yong

    2017-12-01

    This paper proposes a method of accurately assessing image quality without a reference image by using a deep convolutional neural network. Existing training based methods usually utilize a compact set of linear filters for learning features of images captured by different sensors to assess their quality. These methods may not be able to learn the semantic features that are intimately related with the features used in human subject assessment. Observing this drawback, this work proposes training a deep convolutional neural network (CNN) with labelled images for image quality assessment. The ReLU in the CNN allows non-linear transformations for extracting high-level image features, providing a more reliable assessment of image quality than linear filters. To enable the neural network to take images of any arbitrary size as input, the spatial pyramid pooling (SPP) is introduced connecting the top convolutional layer and the fully-connected layer. In addition, the SPP makes the CNN robust to object deformations to a certain extent. The proposed method taking an image as input carries out an end-to-end learning process, and outputs the quality of the image. It is tested on public datasets. Experimental results show that it outperforms existing methods by a large margin and can accurately assess the image quality on images taken by different sensors of varying sizes.

  18. Adaptive statistical iterative reconstruction use for radiation dose reduction in pediatric lower-extremity CT: impact on diagnostic image quality.

    PubMed

    Shah, Amisha; Rees, Mitchell; Kar, Erica; Bolton, Kimberly; Lee, Vincent; Panigrahy, Ashok

    2018-06-01

    For the past several years, increased levels of imaging radiation and cumulative radiation to children has been a significant concern. Although several measures have been taken to reduce radiation dose during computed tomography (CT) scan, the newer dose reduction software adaptive statistical iterative reconstruction (ASIR) has been an effective technique in reducing radiation dose. To our knowledge, no studies are published that assess the effect of ASIR on extremity CT scans in children. To compare radiation dose, image noise, and subjective image quality in pediatric lower extremity CT scans acquired with and without ASIR. The study group consisted of 53 patients imaged on a CT scanner equipped with ASIR software. The control group consisted of 37 patients whose CT images were acquired without ASIR. Image noise, Computed Tomography Dose Index (CTDI) and dose length product (DLP) were measured. Two pediatric radiologists rated the studies in subjective categories: image sharpness, noise, diagnostic acceptability, and artifacts. The CTDI (p value = 0.0184) and DLP (p value <0.0002) were significantly decreased with the use of ASIR compared with non-ASIR studies. However, the subjective ratings for sharpness (p < 0.0001) and diagnostic acceptability of the ASIR images (p < 0.0128) were decreased compared with standard, non-ASIR CT studies. Adaptive statistical iterative reconstruction reduces radiation dose for lower extremity CTs in children, but at the expense of diagnostic imaging quality. Further studies are warranted to determine the specific utility of ASIR for pediatric musculoskeletal CT imaging.

  19. The study of surgical image quality evaluation system by subjective quality factor method

    NASA Astrophysics Data System (ADS)

    Zhang, Jian J.; Xuan, Jason R.; Yang, Xirong; Yu, Honggang; Koullick, Edouard

    2016-03-01

    GreenLightTM procedure is an effective and economical way of treatment of benign prostate hyperplasia (BPH); there are almost a million of patients treated with GreenLightTM worldwide. During the surgical procedure, the surgeon or physician will rely on the monitoring video system to survey and confirm the surgical progress. There are a few obstructions that could greatly affect the image quality of the monitoring video, like laser glare by the tissue and body fluid, air bubbles and debris generated by tissue evaporation, and bleeding, just to name a few. In order to improve the physician's visual experience of a laser surgical procedure, the system performance parameter related to image quality needs to be well defined. However, since image quality is the integrated set of perceptions of the overall degree of excellence of an image, or in other words, image quality is the perceptually weighted combination of significant attributes (contrast, graininess …) of an image when considered in its marketplace or application, there is no standard definition on overall image or video quality especially for the no-reference case (without a standard chart as reference). In this study, Subjective Quality Factor (SQF) and acutance are used for no-reference image quality evaluation. Basic image quality parameters, like sharpness, color accuracy, size of obstruction and transmission of obstruction, are used as subparameter to define the rating scale for image quality evaluation or comparison. Sample image groups were evaluated by human observers according to the rating scale. Surveys of physician groups were also conducted with lab generated sample videos. The study shows that human subjective perception is a trustworthy way of image quality evaluation. More systematic investigation on the relationship between video quality and image quality of each frame will be conducted as a future study.

  20. Study of Image Quality From CT Scanner Multi-Detector by using Americans College of Radiology (ACR) Phantom

    NASA Astrophysics Data System (ADS)

    Mulyadin; Dewang, Syamsir; Abdullah, Bualkar; Tahir, Dahlang

    2018-03-01

    In this study, the image quality of CT scan using phantom American College of Radiology (ACR) was determined. Scanning multidetector CT is used to know the image quality parameters by using a solid phantom containing four modules and primarily from materials that are equivalent to water. Each module is 4 cm in diameter and 20 cm in diameter. There is white alignment marks painted white to reflect the alignment laser and there are also “HEAD”, “FOOT”, and “TOP” marks on the phantom to help align. This test obtains CT images of each module according to the routine inspection protocol of the head. Acceptance of image quality obtained for determination: CT Number Accuracy (CTN), CT Number Uniformity and Noise, Linearity CT Number, Slice Technique, Low Contrast Resolution and High Contrast Resolution represent image quality parameters. In testing CT Number Accuracy (CTN), CT Uniform number and Noise are in the range of tolerable values allowed. In the test, Linearity CT Number obtained correlation value above 0.99 is the relationship between electron density and CT Number. In a low contrast resolution test, the smallest contrast groups are visible. In contrast, the high resolution is seen up to 7 lp/cm. The quality of GE CT Scan is very high, as all the image quality tests obtained are within the tolerance brackets of values permitted by the Nuclear Power Control Agency (BAPETEN). Image quality test is a way to get very important information about the accuracy of snoring result by using phantom ACR.

  1. Modified-BRISQUE as no reference image quality assessment for structural MR images.

    PubMed

    Chow, Li Sze; Rajagopal, Heshalini

    2017-11-01

    An effective and practical Image Quality Assessment (IQA) model is needed to assess the image quality produced from any new hardware or software in MRI. A highly competitive No Reference - IQA (NR - IQA) model called Blind/Referenceless Image Spatial Quality Evaluator (BRISQUE) initially designed for natural images were modified to evaluate structural MR images. The BRISQUE model measures the image quality by using the locally normalized luminance coefficients, which were used to calculate the image features. The modified-BRISQUE model trained a new regression model using MR image features and Difference Mean Opinion Score (DMOS) from 775 MR images. Two types of benchmarks: objective and subjective assessments were used as performance evaluators for both original and modified-BRISQUE models. There was a high correlation between the modified-BRISQUE with both benchmarks, and they were higher than those for the original BRISQUE. There was a significant percentage improvement in their correlation values. The modified-BRISQUE was statistically better than the original BRISQUE. The modified-BRISQUE model can accurately measure the image quality of MR images. It is a practical NR-IQA model for MR images without using reference images. Copyright © 2017 Elsevier Inc. All rights reserved.

  2. Quantitative image quality evaluation of MR images using perceptual difference models

    PubMed Central

    Miao, Jun; Huo, Donglai; Wilson, David L.

    2008-01-01

    The authors are using a perceptual difference model (Case-PDM) to quantitatively evaluate image quality of the thousands of test images which can be created when optimizing fast magnetic resonance (MR) imaging strategies and reconstruction techniques. In this validation study, they compared human evaluation of MR images from multiple organs and from multiple image reconstruction algorithms to Case-PDM and similar models. The authors found that Case-PDM compared very favorably to human observers in double-stimulus continuous-quality scale and functional measurement theory studies over a large range of image quality. The Case-PDM threshold for nonperceptible differences in a 2-alternative forced choice study varied with the type of image under study, but was ≈1.1 for diffuse image effects, providing a rule of thumb. Ordering the image quality evaluation models, we found in overall Case-PDM ≈ IDM (Sarnoff Corporation) ≈ SSIM [Wang et al. IEEE Trans. Image Process. 13, 600–612 (2004)] > mean squared error ≈ NR [Wang et al. (2004) (unpublished)] > DCTune (NASA) > IQM (MITRE Corporation). The authors conclude that Case-PDM is very useful in MR image evaluation but that one should probably restrict studies to similar images and similar processing, normally not a limitation in image reconstruction studies. PMID:18649487

  3. Objective criteria for acceptability and constancy tests of digital subtraction angiography.

    PubMed

    de las Heras, Hugo; Torres, Ricardo; Fernández-Soto, José Miguel; Vañó, Eliseo

    2016-01-01

    Demonstrate an objective procedure to quantify image quality in digital subtraction angiography (DSA) and suggest thresholds for acceptability and constancy tests. Series of images were obtained in a DSA system simulating a small (paediatric) and a large patient using the dynamic phantom described in the IEC and DIN standards for acceptance tests of DSA equipment. Image quality was quantified using measurements of contrast-to-noise ratio (CNR). Overall scores combining the CNR of 10-100 mg/ml Iodine at a vascular diameter of 1-4 mm in a homogeneous background were defined. Phantom entrance surface air kerma (Ka,e) was measured with an ionisation chamber. The visibility of a low-contrast vessel in DSA images has been identified with a CNR value of 0.50 ± 0.03. Despite using 14 times more Ka,e (8.85 vs 0.63 mGy/image), the protocol for large patients showed a decrease in the overall score CNRsum of 67% (4.21 ± 0.06 vs 2.10 ± 0.05). The uncertainty in the results of the objective method was below 5%. Objective evaluation of DSA images using CNR is feasible with dedicated phantom measurements. An objective methodology has been suggested for acceptance tests compliant with the IEC/DIN standards. The defined overall scores can serve to fix a reproducible baseline for constancy tests, as well as to study the device stability within one acquisition series and compare different imaging protocols. This work provides aspects that have not been included in the recent European guidelines on Criteria for Acceptability of Medical Radiological Equipment. Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  4. Low tube voltage computed tomography urography using low-concentration contrast media: Comparison of image quality in conventional computed tomography urography.

    PubMed

    Hwang, Inpyeong; Cho, Jeong Yeon; Kim, Sang Youn; Oh, Seung-June; Ku, Ja Hyeon; Lee, Joongyup; Kim, Seung Hyup

    2015-12-01

    The aim of the present study was to investigate the feasibility and image quality of excretory CT urography performed using low iodine-concentration contrast media and low tube voltage. This prospective study enrolled 63 patients who undergoing CT urography. The subjects were randomized into two groups of an excretory phase CT urography protocol and received either 240 mg I/mL of contrast media and 80 kVp of tube voltage (low-concentration protocol, n=32) or 350 mg I/mL and 120 kVp (conventional protocol, n=31). Two readers qualitatively evaluated images for sharpness of the urinary tract, image noise, streak artifact and overall diagnostic acceptability. The mean attenuation, signal-to-noise ratio, contrast-to-noise ratio and figure of merit were measured in the urinary tract. The non-inferiority test assessed the diagnostic acceptability between the two protocol groups. The low-concentration protocol showed a significantly lower effective radiation dose (3.44 vs. 5.70 mSv, P<.001). The diagnostic acceptability was significantly lower in the low-concentration protocol with iterative reconstruction algorithm than in the conventional protocol (4.06±0.45 vs. 4.50±0.37, P<.001), however, all subjects showed at least more than standard diagnostic acceptability and the difference resided in the predefined non-inferiority margin. The signal-to-noise ratio, contrast-to-noise ratio and figure of merit were significantly higher in the low-concentration protocol along the entire urinary tract (P<.001). CT urography using 240 mg I/mL iodine contrast media, 80 kVp tube voltage and an iterative reconstruction algorithm is beneficial to reduce radiation dose and iodine load, and its objective image quality and subjective diagnostic acceptability is not inferior to that of conventional CT urography. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  5. Quality-control issues on high-resolution diagnostic monitors.

    PubMed

    Parr, L F; Anderson, A L; Glennon, B K; Fetherston, P

    2001-06-01

    Previous literature indicates a need for more data collection in the area of quality control of high-resolution diagnostic monitors. Throughout acceptance testing, which began in June 2000, stability of monitor calibration was analyzed. Although image quality on all monitors was found to be acceptable upon initial acceptance testing using VeriLUM software by Image Smiths, Inc (Germantown, MD), it was determined to be unacceptable during the clinical phase of acceptance testing. High-resolution monitors were evaluated for quality assurance on a weekly basis from installation through acceptance testing and beyond. During clinical utilization determination (CUD), monitor calibration was identified as a problem and the manufacturer returned and recalibrated all workstations. From that time through final acceptance testing, high-resolution monitor calibration and monitor failure rate remained a problem. The monitor vendor then returned to the site to address these areas. Monitor defocus was still noticeable and calibration checks were increased to three times per week. White and black level drift on medium-resolution monitors had been attributed to raster size settings. Measurements of white and black level at several different size settings were taken to determine the effect of size on white and black level settings. Black level remained steady with size change. White level appeared to increase by 2.0 cd/m2 for every 0.1 inches decrease in horizontal raster size. This was determined not to be the cause of the observed brightness drift. Frequency of calibration/testing is an issue in a clinical environment. The increased frequency required at our site cannot be sustained. The medical physics division cannot provide dedicated personnel to conduct the quality-assurance testing on all monitors at this interval due to other physics commitments throughout the hospital. Monitor access is also an issue due to radiologists' need to read images. Some workstations are in use 7 AM

  6. Retinal image quality assessment based on image clarity and content

    NASA Astrophysics Data System (ADS)

    Abdel-Hamid, Lamiaa; El-Rafei, Ahmed; El-Ramly, Salwa; Michelson, Georg; Hornegger, Joachim

    2016-09-01

    Retinal image quality assessment (RIQA) is an essential step in automated screening systems to avoid misdiagnosis caused by processing poor quality retinal images. A no-reference transform-based RIQA algorithm is introduced that assesses images based on five clarity and content quality issues: sharpness, illumination, homogeneity, field definition, and content. Transform-based RIQA algorithms have the advantage of considering retinal structures while being computationally inexpensive. Wavelet-based features are proposed to evaluate the sharpness and overall illumination of the images. A retinal saturation channel is designed and used along with wavelet-based features for homogeneity assessment. The presented sharpness and illumination features are utilized to assure adequate field definition, whereas color information is used to exclude nonretinal images. Several publicly available datasets of varying quality grades are utilized to evaluate the feature sets resulting in area under the receiver operating characteristic curve above 0.99 for each of the individual feature sets. The overall quality is assessed by a classifier that uses the collective features as an input vector. The classification results show superior performance of the algorithm in comparison to other methods from literature. Moreover, the algorithm addresses efficiently and comprehensively various quality issues and is suitable for automatic screening systems.

  7. Remote Sensing Image Quality Assessment Experiment with Post-Processing

    NASA Astrophysics Data System (ADS)

    Jiang, W.; Chen, S.; Wang, X.; Huang, Q.; Shi, H.; Man, Y.

    2018-04-01

    This paper briefly describes the post-processing influence assessment experiment, the experiment includes three steps: the physical simulation, image processing, and image quality assessment. The physical simulation models sampled imaging system in laboratory, the imaging system parameters are tested, the digital image serving as image processing input are produced by this imaging system with the same imaging system parameters. The gathered optical sampled images with the tested imaging parameters are processed by 3 digital image processes, including calibration pre-processing, lossy compression with different compression ratio and image post-processing with different core. Image quality assessment method used is just noticeable difference (JND) subject assessment based on ISO20462, through subject assessment of the gathered and processing images, the influence of different imaging parameters and post-processing to image quality can be found. The six JND subject assessment experimental data can be validated each other. Main conclusions include: image post-processing can improve image quality; image post-processing can improve image quality even with lossy compression, image quality with higher compression ratio improves less than lower ratio; with our image post-processing method, image quality is better, when camera MTF being within a small range.

  8. Subjective matters: from image quality to image psychology

    NASA Astrophysics Data System (ADS)

    Fedorovskaya, Elena A.; De Ridder, Huib

    2013-03-01

    From the advent of digital imaging through several decades of studies, the human vision research community systematically focused on perceived image quality and digital artifacts due to resolution, compression, gamma, dynamic range, capture and reproduction noise, blur, etc., to help overcome existing technological challenges and shortcomings. Technological advances made digital images and digital multimedia nearly flawless in quality, and ubiquitous and pervasive in usage, provide us with the exciting but at the same time demanding possibility to turn to the domain of human experience including higher psychological functions, such as cognition, emotion, awareness, social interaction, consciousness and Self. In this paper we will outline the evolution of human centered multidisciplinary studies related to imaging and propose steps and potential foci of future research.

  9. Feature maps driven no-reference image quality prediction of authentically distorted images

    NASA Astrophysics Data System (ADS)

    Ghadiyaram, Deepti; Bovik, Alan C.

    2015-03-01

    Current blind image quality prediction models rely on benchmark databases comprised of singly and synthetically distorted images, thereby learning image features that are only adequate to predict human perceived visual quality on such inauthentic distortions. However, real world images often contain complex mixtures of multiple distortions. Rather than a) discounting the effect of these mixtures of distortions on an image's perceptual quality and considering only the dominant distortion or b) using features that are only proven to be efficient for singly distorted images, we deeply study the natural scene statistics of authentically distorted images, in different color spaces and transform domains. We propose a feature-maps-driven statistical approach which avoids any latent assumptions about the type of distortion(s) contained in an image, and focuses instead on modeling the remarkable consistencies in the scene statistics of real world images in the absence of distortions. We design a deep belief network that takes model-based statistical image features derived from a very large database of authentically distorted images as input and discovers good feature representations by generalizing over different distortion types, mixtures, and severities, which are later used to learn a regressor for quality prediction. We demonstrate the remarkable competence of our features for improving automatic perceptual quality prediction on a benchmark database and on the newly designed LIVE Authentic Image Quality Challenge Database and show that our approach of combining robust statistical features and the deep belief network dramatically outperforms the state-of-the-art.

  10. TU-H-CAMPUS-JeP1-04: Deformable Image Registration Performances in Pelvis Patients: Impact of CBCT Image Quality

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fusella, M; Loi, G; Fiandra, C

    Purpose: To investigate the accuracy and robustness, against image noise and artifacts (typical of CBCT images), of a commercial algorithm for deformable image registration (DIR), to propagate regions of interest (ROIs) in computational phantoms based on real prostate patient images. Methods: The Anaconda DIR algorithm, implemented in RayStation was tested. Two specific Deformation Vector Fields (DVFs) were applied to the reference data set (CTref) using the ImSimQA software, obtaining two deformed CTs. For each dataset twenty-four different level of noise and/or capping artifacts were applied to simulate CBCT images. DIR was performed between CTref and each deformed CTs and CBCTs.more » In order to investigate the relationship between image quality parameters and the DIR results (expressed by a logit transform of the Dice Index) a bilinear regression was defined. Results: More than 550 DIR-mapped ROIs were analyzed. The Statistical analysis states that deformation strenght and artifacts were significant prognostic factors of DIR performances, while noise appeared to have a minor role in DIR process as implemented in RayStation as expected by the image similarity metric built in the registration algorithm. Capping artifacts reveals a determinant role for the accuracy of DIR results. Two optimal values for capping artifacts were found to obtain acceptable DIR results (DICE> 075/ 0.85). Various clinical CBCT acquisition protocol were reported to evaluate the significance of the study. Conclusion: This work illustrates the impact of image quality on DIR performance. Clinical issues like Adaptive Radiation Therapy (ART) and Dose Accumulation need accurate and robust DIR software. The RayStation DIR algorithm resulted robust against noise, but sensitive to image artifacts. This result highlights the need of robustness quality assurance against image noise and artifacts in the commissioning of a DIR commercial system and underlines the importance to adopt optimized

  11. Objective quality assessment for multiexposure multifocus image fusion.

    PubMed

    Hassen, Rania; Wang, Zhou; Salama, Magdy M A

    2015-09-01

    There has been a growing interest in image fusion technologies, but how to objectively evaluate the quality of fused images has not been fully understood. Here, we propose a method for objective quality assessment of multiexposure multifocus image fusion based on the evaluation of three key factors of fused image quality: 1) contrast preservation; 2) sharpness; and 3) structure preservation. Subjective experiments are conducted to create an image fusion database, based on which, performance evaluation shows that the proposed fusion quality index correlates well with subjective scores, and gives a significant improvement over the existing fusion quality measures.

  12. No-reference image quality assessment for horizontal-path imaging scenarios

    NASA Astrophysics Data System (ADS)

    Rios, Carlos; Gladysz, Szymon

    2013-05-01

    There exist several image-enhancement algorithms and tasks associated with imaging through turbulence that depend on defining the quality of an image. Examples include: "lucky imaging", choosing the width of the inverse filter for image reconstruction, or stopping iterative deconvolution. We collected a number of image quality metrics found in the literature. Particularly interesting are the blind, "no-reference" metrics. We discuss ways of evaluating the usefulness of these metrics, even when a fully objective comparison is impossible because of the lack of a reference image. Metrics are tested on simulated and real data. Field data comes from experiments performed by the NATO SET 165 research group over a 7 km distance in Dayton, Ohio.

  13. Guidance for Efficient Small Animal Imaging Quality Control.

    PubMed

    Osborne, Dustin R; Kuntner, Claudia; Berr, Stuart; Stout, David

    2017-08-01

    Routine quality control is a critical aspect of properly maintaining high-performance small animal imaging instrumentation. A robust quality control program helps produce more reliable data both for academic purposes and as proof of system performance for contract imaging work. For preclinical imaging laboratories, the combination of costs and available resources often limits their ability to produce efficient and effective quality control programs. This work presents a series of simplified quality control procedures that are accessible to a wide range of preclinical imaging laboratories. Our intent is to provide minimum guidelines for routine quality control that can assist preclinical imaging specialists in setting up an appropriate quality control program for their facility.

  14. The effect of image sharpness on quantitative eye movement data and on image quality evaluation while viewing natural images

    NASA Astrophysics Data System (ADS)

    Vuori, Tero; Olkkonen, Maria

    2006-01-01

    The aim of the study is to test both customer image quality rating (subjective image quality) and physical measurement of user behavior (eye movements tracking) to find customer satisfaction differences in imaging technologies. Methodological aim is to find out whether eye movements could be quantitatively used in image quality preference studies. In general, we want to map objective or physically measurable image quality to subjective evaluations and eye movement data. We conducted a series of image quality tests, in which the test subjects evaluated image quality while we recorded their eye movements. Results show that eye movement parameters consistently change according to the instructions given to the user, and according to physical image quality, e.g. saccade duration increased with increasing blur. Results indicate that eye movement tracking could be used to differentiate image quality evaluation strategies that the users have. Results also show that eye movements would help mapping between technological and subjective image quality. Furthermore, these results give some empirical emphasis to top-down perception processes in image quality perception and evaluation by showing differences between perceptual processes in situations when cognitive task varies.

  15. Global quality imaging: improvement actions.

    PubMed

    Lau, Lawrence S; Pérez, Maria R; Applegate, Kimberly E; Rehani, Madan M; Ringertz, Hans G; George, Robert

    2011-05-01

    Workforce shortage, workload increase, workplace changes, and budget challenges are emerging issues around the world, which could place quality imaging at risk. It is important for imaging stakeholders to collaborate, ensure patient safety, improve the quality of care, and address these issues. There is no single panacea. A range of improvement measures, strategies, and actions are required. Examples of improvement actions supporting the 3 quality measures are described under 5 strategies: conducting research, promoting awareness, providing education and training, strengthening infrastructure, and implementing policies. The challenge is to develop long-term, cost-effective, system-based improvement actions that will bring better outcomes and underpin a sustainable future for quality imaging. In an imaging practice, these actions will result in selecting the right procedure (justification), using the right dose (optimization), and preventing errors along the patient journey. To realize this vision and implement these improvement actions, a range of expertise and adequate resources are required. Stakeholders should collaborate and work together. In today's globalized environment, collaboration is strength and provides synergy to achieve better outcomes and greater success. Copyright © 2011 American College of Radiology. Published by Elsevier Inc. All rights reserved.

  16. Statistical iterative reconstruction to improve image quality for digital breast tomosynthesis

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Xu, Shiyu, E-mail: shiyu.xu@gmail.com; Chen, Ying, E-mail: adachen@siu.edu; Lu, Jianping

    2015-09-15

    Purpose: Digital breast tomosynthesis (DBT) is a novel modality with the potential to improve early detection of breast cancer by providing three-dimensional (3D) imaging with a low radiation dose. 3D image reconstruction presents some challenges: cone-beam and flat-panel geometry, and highly incomplete sampling. A promising means to overcome these challenges is statistical iterative reconstruction (IR), since it provides the flexibility of accurate physics modeling and a general description of system geometry. The authors’ goal was to develop techniques for applying statistical IR to tomosynthesis imaging data. Methods: These techniques include the following: a physics model with a local voxel-pair basedmore » prior with flexible parameters to fine-tune image quality; a precomputed parameter λ in the prior, to remove data dependence and to achieve a uniform resolution property; an effective ray-driven technique to compute the forward and backprojection; and an oversampled, ray-driven method to perform high resolution reconstruction with a practical region-of-interest technique. To assess the performance of these techniques, the authors acquired phantom data on the stationary DBT prototype system. To solve the estimation problem, the authors proposed an optimization-transfer based algorithm framework that potentially allows fewer iterations to achieve an acceptably converged reconstruction. Results: IR improved the detectability of low-contrast and small microcalcifications, reduced cross-plane artifacts, improved spatial resolution, and lowered noise in reconstructed images. Conclusions: Although the computational load remains a significant challenge for practical development, the superior image quality provided by statistical IR, combined with advancing computational techniques, may bring benefits to screening, diagnostics, and intraoperative imaging in clinical applications.« less

  17. Quality by design for herbal drugs: a feedforward control strategy and an approach to define the acceptable ranges of critical quality attributes.

    PubMed

    Yan, Binjun; Li, Yao; Guo, Zhengtai; Qu, Haibin

    2014-01-01

    The concept of quality by design (QbD) has been widely accepted and applied in the pharmaceutical manufacturing industry. There are still two key issues to be addressed in the implementation of QbD for herbal drugs. The first issue is the quality variation of herbal raw materials and the second issue is the difficulty in defining the acceptable ranges of critical quality attributes (CQAs). To propose a feedforward control strategy and a method for defining the acceptable ranges of CQAs for the two issues. In the case study of the ethanol precipitation process of Danshen (Radix Salvia miltiorrhiza) injection, regression models linking input material attributes and process parameters to CQAs were built first and an optimisation model for calculating the best process parameters according to the input materials was established. Then, the feasible material space was defined and the acceptable ranges of CQAs for the previous process were determined. In the case study, satisfactory regression models were built with cross-validated regression coefficients (Q(2) ) all above 91 %. The feedforward control strategy was applied successfully to compensate the quality variation of the input materials, which was able to control the CQAs in the 90-110 % ranges of the desired values. In addition, the feasible material space for the ethanol precipitation process was built successfully, which showed the acceptable ranges of the CQAs for the concentration process. The proposed methodology can help to promote the implementation of QbD for herbal drugs. Copyright © 2013 John Wiley & Sons, Ltd.

  18. The use of adaptive statistical iterative reconstruction (ASiR) technique in evaluation of patients with cervical spine trauma: impact on radiation dose reduction and image quality

    PubMed Central

    Sheikh, Adnan

    2016-01-01

    Objective: The aim of this study was to evaluate the impact of adaptive statistical iterative reconstruction (ASiR) technique on the image quality and radiation dose reduction. The comparison was made with the traditional filtered back projection (FBP) technique. Methods: We retrospectively reviewed 78 patients, who underwent cervical spine CT for blunt cervical trauma between 1 June 2010 and 30 November 2010. 48 patients were imaged using traditional FBP technique and the remaining 30 patients were imaged using the ASiR technique. The patient demographics, radiation dose, objective image signal and noise were recorded; while subjective noise, sharpness, diagnostic acceptability and artefacts were graded by two radiologists blinded to the techniques. Results: We found that the ASiR technique was able to reduce the volume CT dose index, dose–length product and effective dose by 36%, 36.5% and 36.5%, respectively, compared with the FBP technique. There was no significant difference in the image noise (p = 0.39), signal (p = 0.82) and signal-to-noise ratio (p = 0.56) between the groups. The subjective image quality was minimally better in the ASiR group but not statistically significant. There was excellent interobserver agreement on the subjective image quality and diagnostic acceptability for both groups. Conclusion: The use of ASiR technique allowed approximately 36% radiation dose reduction in the evaluation of cervical spine without degrading the image quality. Advances in knowledge: The present study highlights that the ASiR technique is extremely helpful in reducing the patient radiation exposure while maintaining the image quality. It is highly recommended to utilize this novel technique in CT imaging of different body regions. PMID:26882825

  19. The use of adaptive statistical iterative reconstruction (ASiR) technique in evaluation of patients with cervical spine trauma: impact on radiation dose reduction and image quality.

    PubMed

    Patro, Satya N; Chakraborty, Santanu; Sheikh, Adnan

    2016-01-01

    The aim of this study was to evaluate the impact of adaptive statistical iterative reconstruction (ASiR) technique on the image quality and radiation dose reduction. The comparison was made with the traditional filtered back projection (FBP) technique. We retrospectively reviewed 78 patients, who underwent cervical spine CT for blunt cervical trauma between 1 June 2010 and 30 November 2010. 48 patients were imaged using traditional FBP technique and the remaining 30 patients were imaged using the ASiR technique. The patient demographics, radiation dose, objective image signal and noise were recorded; while subjective noise, sharpness, diagnostic acceptability and artefacts were graded by two radiologists blinded to the techniques. We found that the ASiR technique was able to reduce the volume CT dose index, dose-length product and effective dose by 36%, 36.5% and 36.5%, respectively, compared with the FBP technique. There was no significant difference in the image noise (p = 0.39), signal (p = 0.82) and signal-to-noise ratio (p = 0.56) between the groups. The subjective image quality was minimally better in the ASiR group but not statistically significant. There was excellent interobserver agreement on the subjective image quality and diagnostic acceptability for both groups. The use of ASiR technique allowed approximately 36% radiation dose reduction in the evaluation of cervical spine without degrading the image quality. The present study highlights that the ASiR technique is extremely helpful in reducing the patient radiation exposure while maintaining the image quality. It is highly recommended to utilize this novel technique in CT imaging of different body regions.

  20. Objective quality assessment of tone-mapped images.

    PubMed

    Yeganeh, Hojatollah; Wang, Zhou

    2013-02-01

    Tone-mapping operators (TMOs) that convert high dynamic range (HDR) to low dynamic range (LDR) images provide practically useful tools for the visualization of HDR images on standard LDR displays. Different TMOs create different tone-mapped images, and a natural question is which one has the best quality. Without an appropriate quality measure, different TMOs cannot be compared, and further improvement is directionless. Subjective rating may be a reliable evaluation method, but it is expensive and time consuming, and more importantly, is difficult to be embedded into optimization frameworks. Here we propose an objective quality assessment algorithm for tone-mapped images by combining: 1) a multiscale signal fidelity measure on the basis of a modified structural similarity index and 2) a naturalness measure on the basis of intensity statistics of natural images. Validations using independent subject-rated image databases show good correlations between subjective ranking score and the proposed tone-mapped image quality index (TMQI). Furthermore, we demonstrate the extended applications of TMQI using two examples-parameter tuning for TMOs and adaptive fusion of multiple tone-mapped images.

  1. Low-cost oblique illumination: an image quality assessment.

    PubMed

    Ruiz-Santaquiteria, Jesus; Espinosa-Aranda, Jose Luis; Deniz, Oscar; Sanchez, Carlos; Borrego-Ramos, Maria; Blanco, Saul; Cristobal, Gabriel; Bueno, Gloria

    2018-01-01

    We study the effectiveness of several low-cost oblique illumination filters to improve overall image quality, in comparison with standard bright field imaging. For this purpose, a dataset composed of 3360 diatom images belonging to 21 taxa was acquired. Subjective and objective image quality assessments were done. The subjective evaluation was performed by a group of diatom experts by psychophysical test where resolution, focus, and contrast were assessed. Moreover, some objective nonreference image quality metrics were applied to the same image dataset to complete the study, together with the calculation of several texture features to analyze the effect of these filters in terms of textural properties. Both image quality evaluation methods, subjective and objective, showed better results for images acquired using these illumination filters in comparison with the no filtered image. These promising results confirm that this kind of illumination filters can be a practical way to improve the image quality, thanks to the simple and low cost of the design and manufacturing process. (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE).

  2. Quality metrics for sensor images

    NASA Technical Reports Server (NTRS)

    Ahumada, AL

    1993-01-01

    Methods are needed for evaluating the quality of augmented visual displays (AVID). Computational quality metrics will help summarize, interpolate, and extrapolate the results of human performance tests with displays. The FLM Vision group at NASA Ames has been developing computational models of visual processing and using them to develop computational metrics for similar problems. For example, display modeling systems use metrics for comparing proposed displays, halftoning optimizing methods use metrics to evaluate the difference between the halftone and the original, and image compression methods minimize the predicted visibility of compression artifacts. The visual discrimination models take as input two arbitrary images A and B and compute an estimate of the probability that a human observer will report that A is different from B. If A is an image that one desires to display and B is the actual displayed image, such an estimate can be regarded as an image quality metric reflecting how well B approximates A. There are additional complexities associated with the problem of evaluating the quality of radar and IR enhanced displays for AVID tasks. One important problem is the question of whether intruding obstacles are detectable in such displays. Although the discrimination model can handle detection situations by making B the original image A plus the intrusion, this detection model makes the inappropriate assumption that the observer knows where the intrusion will be. Effects of signal uncertainty need to be added to our models. A pilot needs to make decisions rapidly. The models need to predict not just the probability of a correct decision, but the probability of a correct decision by the time the decision needs to be made. That is, the models need to predict latency as well as accuracy. Luce and Green have generated models for auditory detection latencies. Similar models are needed for visual detection. Most image quality models are designed for static imagery

  3. A Dynamic Image Quality Evaluation of Videofluoroscopy Images: Considerations for Telepractice Applications.

    PubMed

    Burns, Clare L; Keir, Benjamin; Ward, Elizabeth C; Hill, Anne J; Farrell, Anna; Phillips, Nick; Porter, Linda

    2015-08-01

    High-quality fluoroscopy images are required for accurate interpretation of videofluoroscopic swallow studies (VFSS) by speech pathologists and radiologists. Consequently, integral to developing any system to conduct VFSS remotely via telepractice is ensuring that the quality of the VFSS images transferred via the telepractice system is optimized. This study evaluates the extent of change observed in image quality when videofluoroscopic images are transmitted from a digital fluoroscopy system to (a) current clinical equipment (KayPentax Digital Swallowing Workstation, and b) four different telepractice system configurations. The telepractice system configurations consisted of either a local C20 or C60 Cisco TelePresence System (codec unit) connected to the digital fluoroscopy system and linked to a second remote C20 or C60 Cisco TelePresence System via a network running at speeds of either 2, 4 or 6 megabits per second (Mbit/s). Image quality was tested using the NEMA XR 21 Phantom, and results demonstrated some loss in spatial resolution, low contrast detectability and temporal resolution for all transferred images when compared to the fluoroscopy source. When using higher capacity codec units and/or the highest bandwidths to support data transmission, image quality transmitted through the telepractice system was found to be comparable if not better than the current clinical system. This study confirms that telepractice systems can be designed to support fluoroscopy image transfer and highlights important considerations when developing telepractice systems for VFSS analysis to ensure high-quality radiological image reproduction.

  4. Learning to rank for blind image quality assessment.

    PubMed

    Gao, Fei; Tao, Dacheng; Gao, Xinbo; Li, Xuelong

    2015-10-01

    Blind image quality assessment (BIQA) aims to predict perceptual image quality scores without access to reference images. State-of-the-art BIQA methods typically require subjects to score a large number of images to train a robust model. However, subjective quality scores are imprecise, biased, and inconsistent, and it is challenging to obtain a large-scale database, or to extend existing databases, because of the inconvenience of collecting images, training the subjects, conducting subjective experiments, and realigning human quality evaluations. To combat these limitations, this paper explores and exploits preference image pairs (PIPs) such as the quality of image Ia is better than that of image Ib for training a robust BIQA model. The preference label, representing the relative quality of two images, is generally precise and consistent, and is not sensitive to image content, distortion type, or subject identity; such PIPs can be generated at a very low cost. The proposed BIQA method is one of learning to rank. We first formulate the problem of learning the mapping from the image features to the preference label as one of classification. In particular, we investigate the utilization of a multiple kernel learning algorithm based on group lasso to provide a solution. A simple but effective strategy to estimate perceptual image quality scores is then presented. Experiments show that the proposed BIQA method is highly effective and achieves a performance comparable with that of state-of-the-art BIQA algorithms. Moreover, the proposed method can be easily extended to new distortion categories.

  5. Adaptive Statistical Iterative Reconstruction-V Versus Adaptive Statistical Iterative Reconstruction: Impact on Dose Reduction and Image Quality in Body Computed Tomography.

    PubMed

    Gatti, Marco; Marchisio, Filippo; Fronda, Marco; Rampado, Osvaldo; Faletti, Riccardo; Bergamasco, Laura; Ropolo, Roberto; Fonio, Paolo

    The aim of this study was to evaluate the impact on dose reduction and image quality of the new iterative reconstruction technique: adaptive statistical iterative reconstruction (ASIR-V). Fifty consecutive oncologic patients acted as case controls undergoing during their follow-up a computed tomography scan both with ASIR and ASIR-V. Each study was analyzed in a double-blinded fashion by 2 radiologists. Both quantitative and qualitative analyses of image quality were conducted. Computed tomography scanner radiation output was 38% (29%-45%) lower (P < 0.0001) for the ASIR-V examinations than for the ASIR ones. The quantitative image noise was significantly lower (P < 0.0001) for ASIR-V. Adaptive statistical iterative reconstruction-V had a higher performance for the subjective image noise (P = 0.01 for 5 mm and P = 0.009 for 1.25 mm), the other parameters (image sharpness, diagnostic acceptability, and overall image quality) being similar (P > 0.05). Adaptive statistical iterative reconstruction-V is a new iterative reconstruction technique that has the potential to provide image quality equal to or greater than ASIR, with a dose reduction around 40%.

  6. Standardizing Quality Assessment of Fused Remotely Sensed Images

    NASA Astrophysics Data System (ADS)

    Pohl, C.; Moellmann, J.; Fries, K.

    2017-09-01

    The multitude of available operational remote sensing satellites led to the development of many image fusion techniques to provide high spatial, spectral and temporal resolution images. The comparison of different techniques is necessary to obtain an optimized image for the different applications of remote sensing. There are two approaches in assessing image quality: 1. Quantitatively by visual interpretation and 2. Quantitatively using image quality indices. However an objective comparison is difficult due to the fact that a visual assessment is always subject and a quantitative assessment is done by different criteria. Depending on the criteria and indices the result varies. Therefore it is necessary to standardize both processes (qualitative and quantitative assessment) in order to allow an objective image fusion quality evaluation. Various studies have been conducted at the University of Osnabrueck (UOS) to establish a standardized process to objectively compare fused image quality. First established image fusion quality assessment protocols, i.e. Quality with No Reference (QNR) and Khan's protocol, were compared on varies fusion experiments. Second the process of visual quality assessment was structured and standardized with the aim to provide an evaluation protocol. This manuscript reports on the results of the comparison and provides recommendations for future research.

  7. A qualitative and quantitative analysis of radiation dose and image quality of computed tomography images using adaptive statistical iterative reconstruction

    PubMed Central

    Mail, Noor; Shamy, Abdulrahman M.; Alghamdi, Suliman; Saoudi, Abdelhamid

    2016-01-01

    Image quality is a key issue in radiology, particularly in a clinical setting where it is important to achieve accurate diagnoses while minimizing radiation dose. Some computed tomography (CT) manufacturers have introduced algorithms that claim significant dose reduction. In this study, we assessed CT image quality produced by two reconstruction algorithms provided with GE Healthcare's Discovery 690 Elite positron emission tomography (PET) CT scanner. Image quality was measured for images obtained at various doses with both conventional filtered back‐projection (FBP) and adaptive statistical iterative reconstruction (ASIR) algorithms. A standard CT dose index (CTDI) phantom and a pencil ionization chamber were used to measure the CT dose at 120 kVp and an exposure of 260 mAs. Image quality was assessed using two phantoms. CT images of both phantoms were acquired at tube voltage (kV) of 120 with exposures ranging from 25 mAs to 400 mAs. Images were reconstructed using FBP and ASIR ranging from 10% to 100%, then analyzed for noise, low‐contrast detectability, contrast‐to‐noise ratio (CNR), and modulation transfer function (MTF). Noise was 4.6 HU in water phantom images acquired at 260 mAs/FBP 120 kV and 130 mAs/50% ASIR 120 kV. The large objects (frequency<7 lp/cm) retained fairly acceptable image quality at 130 mAs/50% ASIR, compared to 260 mAs/FBP. The application of ASIR for small objects (frequency>7 lp/cm) showed poor visibility compared to FBP at 260 mAs and even worse for images acquired at less than 130 mAs. ASIR blending more than 50% at low dose tends to reduce contrast of small objects (frequency>7 lp/cm). We concluded that dose reduction and ASIR should be applied with close attention if the objects to be detected or diagnosed are small (frequency>7 lp/cm). Further investigations are required to correlate the small objects (frequency>7 lp/cm) to patient anatomy and clinical diagnosis. PACS number(s): 87.57.‐s, 87.57.C, 87.57.cf, 87

  8. Image quality evaluation of full reference algorithm

    NASA Astrophysics Data System (ADS)

    He, Nannan; Xie, Kai; Li, Tong; Ye, Yushan

    2018-03-01

    Image quality evaluation is a classic research topic, the goal is to design the algorithm, given the subjective feelings consistent with the evaluation value. This paper mainly introduces several typical reference methods of Mean Squared Error(MSE), Peak Signal to Noise Rate(PSNR), Structural Similarity Image Metric(SSIM) and feature similarity(FSIM) of objective evaluation methods. The different evaluation methods are tested by Matlab, and the advantages and disadvantages of these methods are obtained by analyzing and comparing them.MSE and PSNR are simple, but they are not considered to introduce HVS characteristics into image quality evaluation. The evaluation result is not ideal. SSIM has a good correlation and simple calculation ,because it is considered to the human visual effect into image quality evaluation,However the SSIM method is based on a hypothesis,The evaluation result is limited. The FSIM method can be used for test of gray image and color image test, and the result is better. Experimental results show that the new image quality evaluation algorithm based on FSIM is more accurate.

  9. Using the NEMA NU 4 PET image quality phantom in multipinhole small-animal SPECT.

    PubMed

    Harteveld, Anita A; Meeuwis, Antoi P W; Disselhorst, Jonathan A; Slump, Cornelis H; Oyen, Wim J G; Boerman, Otto C; Visser, Eric P

    2011-10-01

    Several commercial small-animal SPECT scanners using multipinhole collimation are presently available. However, generally accepted standards to characterize the performance of these scanners do not exist. Whereas for small-animal PET, the National Electrical Manufacturers Association (NEMA) NU 4 standards have been defined in 2008, such standards are still lacking for small-animal SPECT. In this study, the image quality parameters associated with the NEMA NU 4 image quality phantom were determined for a small-animal multipinhole SPECT scanner. Multiple whole-body scans of the NEMA NU 4 image quality phantom of 1-h duration were performed in a U-SPECT-II scanner using (99m)Tc with activities ranging between 8.4 and 78.2 MBq. The collimator contained 75 pinholes of 1.0-mm diameter and had a bore diameter of 98 mm. Image quality parameters were determined as a function of average phantom activity, number of iterations, postreconstruction spatial filter, and scatter correction. In addition, a mouse was injected with (99m)Tc-hydroxymethylene diphosphonate and was euthanized 6.5 h after injection. Multiple whole-body scans of this mouse of 1-h duration were acquired for activities ranging between 3.29 and 52.7 MBq. An increase in the number of iterations was accompanied by an increase in the recovery coefficients for the small rods (RC(rod)), an increase in the noise in the uniform phantom region, and a decrease in spillover ratios for the cold-air- and water-filled scatter compartments (SOR(air) and SOR(wat)). Application of spatial filtering reduced image noise but lowered RC(rod). Filtering did not influence SOR(air) and SOR(wat). Scatter correction reduced SOR(air) and SOR(wat). The effect of total phantom activity was primarily seen in a reduction of image noise with increasing activity. RC(rod), SOR(air), and SOR(wat) were more or less constant as a function of phantom activity. The relation between acquisition and reconstruction settings and image quality was

  10. High-quality compressive ghost imaging

    NASA Astrophysics Data System (ADS)

    Huang, Heyan; Zhou, Cheng; Tian, Tian; Liu, Dongqi; Song, Lijun

    2018-04-01

    We propose a high-quality compressive ghost imaging method based on projected Landweber regularization and guided filter, which effectively reduce the undersampling noise and improve the resolution. In our scheme, the original object is reconstructed by decomposing of regularization and denoising steps instead of solving a minimization problem in compressive reconstruction process. The simulation and experimental results show that our method can obtain high ghost imaging quality in terms of PSNR and visual observation.

  11. Examples of subjective image quality enhancement in multimedia

    NASA Astrophysics Data System (ADS)

    Klíma, Miloš; Pazderák, Jiří; Fliegel, Karel

    2007-09-01

    The subjective image quality is an important issue in all multimedia imaging systems with a significant impact onto QoS (Quality of Service). For long time the image fidelity criterion was widely applied in technical systems esp. in both television and image source compression fields but the optimization of subjective perception quality and fidelity approach (such as the minimum of MSE) are very different. The paper presents an experimental testing of three different digital techniques for the subjective image quality enhancement - color saturation, edge enhancement, denoising operators and noise addition - well known from both the digital photography and video. The evaluation has been done for extensive operator parameterization and the results are summarized and discussed. It has been demonstrated that there are relevant types of image corrections improving to some extent the subjective perception of the image. The above mentioned techniques have been tested for five image tests with significantly different image characteristics (fine details, large saturated color areas, high color contrast, easy-to-remember colors etc.). The experimental results show the way to optimized use of image enhancing operators. Finally the concept of impressiveness as a new possible expression of subjective quality improvement is presented and discussed.

  12. dipIQ: Blind Image Quality Assessment by Learning-to-Rank Discriminable Image Pairs.

    PubMed

    Ma, Kede; Liu, Wentao; Liu, Tongliang; Wang, Zhou; Tao, Dacheng

    2017-05-26

    Objective assessment of image quality is fundamentally important in many image processing tasks. In this work, we focus on learning blind image quality assessment (BIQA) models which predict the quality of a digital image with no access to its original pristine-quality counterpart as reference. One of the biggest challenges in learning BIQA models is the conflict between the gigantic image space (which is in the dimension of the number of image pixels) and the extremely limited reliable ground truth data for training. Such data are typically collected via subjective testing, which is cumbersome, slow, and expensive. Here we first show that a vast amount of reliable training data in the form of quality-discriminable image pairs (DIP) can be obtained automatically at low cost by exploiting largescale databases with diverse image content. We then learn an opinion-unaware BIQA (OU-BIQA, meaning that no subjective opinions are used for training) model using RankNet, a pairwise learning-to-rank (L2R) algorithm, from millions of DIPs, each associated with a perceptual uncertainty level, leading to a DIP inferred quality (dipIQ) index. Extensive experiments on four benchmark IQA databases demonstrate that dipIQ outperforms state-of-the-art OU-BIQA models. The robustness of dipIQ is also significantly improved as confirmed by the group MAximum Differentiation (gMAD) competition method. Furthermore, we extend the proposed framework by learning models with ListNet (a listwise L2R algorithm) on quality-discriminable image lists (DIL). The resulting DIL Inferred Quality (dilIQ) index achieves an additional performance gain.

  13. Performance evaluation of no-reference image quality metrics for face biometric images

    NASA Astrophysics Data System (ADS)

    Liu, Xinwei; Pedersen, Marius; Charrier, Christophe; Bours, Patrick

    2018-03-01

    The accuracy of face recognition systems is significantly affected by the quality of face sample images. The recent established standardization proposed several important aspects for the assessment of face sample quality. There are many existing no-reference image quality metrics (IQMs) that are able to assess natural image quality by taking into account similar image-based quality attributes as introduced in the standardization. However, whether such metrics can assess face sample quality is rarely considered. We evaluate the performance of 13 selected no-reference IQMs on face biometrics. The experimental results show that several of them can assess face sample quality according to the system performance. We also analyze the strengths and weaknesses of different IQMs as well as why some of them failed to assess face sample quality. Retraining an original IQM by using face database can improve the performance of such a metric. In addition, the contribution of this paper can be used for the evaluation of IQMs on other biometric modalities; furthermore, it can be used for the development of multimodality biometric IQMs.

  14. Reducing Error Rates for Iris Image using higher Contrast in Normalization process

    NASA Astrophysics Data System (ADS)

    Aminu Ghali, Abdulrahman; Jamel, Sapiee; Abubakar Pindar, Zahraddeen; Hasssan Disina, Abdulkadir; Mat Daris, Mustafa

    2017-08-01

    Iris recognition system is the most secured, and faster means of identification and authentication. However, iris recognition system suffers a setback from blurring, low contrast and illumination due to low quality image which compromises the accuracy of the system. The acceptance or rejection rates of verified user depend solely on the quality of the image. In many cases, iris recognition system with low image contrast could falsely accept or reject user. Therefore this paper adopts Histogram Equalization Technique to address the problem of False Rejection Rate (FRR) and False Acceptance Rate (FAR) by enhancing the contrast of the iris image. A histogram equalization technique enhances the image quality and neutralizes the low contrast of the image at normalization stage. The experimental result shows that Histogram Equalization Technique has reduced FRR and FAR compared to the existing techniques.

  15. Achieving quality in cardiovascular imaging: proceedings from the American College of Cardiology-Duke University Medical Center Think Tank on Quality in Cardiovascular Imaging.

    PubMed

    Douglas, Pamela; Iskandrian, Ami E; Krumholz, Harlan M; Gillam, Linda; Hendel, Robert; Jollis, James; Peterson, Eric; Chen, Jersey; Masoudi, Frederick; Mohler, Emile; McNamara, Robert L; Patel, Manesh R; Spertus, John

    2006-11-21

    Cardiovascular imaging has enjoyed both rapid technological advances and sustained growth, yet less attention has been focused on quality than in other areas of cardiovascular medicine. To address this deficit, representatives from cardiovascular imaging societies, private payers, government agencies, the medical imaging industry, and experts in quality measurement met, and this report provides an overview of the discussions. A consensus definition of quality in imaging and a convergence of opinion on quality measures across imaging modalities was achieved and are intended to be the start of a process culminating in the development, dissemination, and adoption of quality measures for all cardiovascular imaging modalities.

  16. Naturalness and interestingness of test images for visual quality evaluation

    NASA Astrophysics Data System (ADS)

    Halonen, Raisa; Westman, Stina; Oittinen, Pirkko

    2011-01-01

    Balanced and representative test images are needed to study perceived visual quality in various application domains. This study investigates naturalness and interestingness as image quality attributes in the context of test images. Taking a top-down approach we aim to find the dimensions which constitute naturalness and interestingness in test images and the relationship between these high-level quality attributes. We compare existing collections of test images (e.g. Sony sRGB images, ISO 12640 images, Kodak images, Nokia images and test images developed within our group) in an experiment combining quality sorting and structured interviews. Based on the data gathered we analyze the viewer-supplied criteria for naturalness and interestingness across image types, quality levels and judges. This study advances our understanding of subjective image quality criteria and enables the validation of current test images, furthering their development.

  17. Quantifying the quality of medical x-ray images: An evaluation based on normal anatomy for lumbar spine and chest radiography

    NASA Astrophysics Data System (ADS)

    Tingberg, Anders Martin

    Optimisation in diagnostic radiology requires accurate methods for determination of patient absorbed dose and clinical image quality. Simple methods for evaluation of clinical image quality are at present scarce and this project aims at developing such methods. Two methods are used and further developed; fulfillment of image criteria (IC) and visual grading analysis (VGA). Clinical image quality descriptors are defined based on these two methods: image criteria score (ICS) and visual grading analysis score (VGAS), respectively. For both methods the basis is the Image Criteria of the ``European Guidelines on Quality Criteria for Diagnostic Radiographic Images''. Both methods have proved to be useful for evaluation of clinical image quality. The two methods complement each other: IC is an absolute method, which means that the quality of images of different patients and produced with different radiographic techniques can be compared with each other. The separating power of IC is, however, weaker than that of VGA. VGA is the best method for comparing images produced with different radiographic techniques and has strong separating power, but the results are relative, since the quality of an image is compared to the quality of a reference image. The usefulness of the two methods has been verified by comparing the results from both of them with results from a generally accepted method for evaluation of clinical image quality, receiver operating characteristics (ROC). The results of the comparison between the two methods based on visibility of anatomical structures and the method based on detection of pathological structures (free-response forced error) indicate that the former two methods can be used for evaluation of clinical image quality as efficiently as the method based on ROC. More studies are, however, needed for us to be able to draw a general conclusion, including studies of other organs, using other radiographic techniques, etc. The results of the experimental

  18. Improving high resolution retinal image quality using speckle illumination HiLo imaging

    PubMed Central

    Zhou, Xiaolin; Bedggood, Phillip; Metha, Andrew

    2014-01-01

    Retinal image quality from flood illumination adaptive optics (AO) ophthalmoscopes is adversely affected by out-of-focus light scatter due to the lack of confocality. This effect is more pronounced in small eyes, such as that of rodents, because the requisite high optical power confers a large dioptric thickness to the retina. A recently-developed structured illumination microscopy (SIM) technique called HiLo imaging has been shown to reduce the effect of out-of-focus light scatter in flood illumination microscopes and produce pseudo-confocal images with significantly improved image quality. In this work, we adopted the HiLo technique to a flood AO ophthalmoscope and performed AO imaging in both (physical) model and live rat eyes. The improvement in image quality from HiLo imaging is shown both qualitatively and quantitatively by using spatial spectral analysis. PMID:25136486

  19. Improving high resolution retinal image quality using speckle illumination HiLo imaging.

    PubMed

    Zhou, Xiaolin; Bedggood, Phillip; Metha, Andrew

    2014-08-01

    Retinal image quality from flood illumination adaptive optics (AO) ophthalmoscopes is adversely affected by out-of-focus light scatter due to the lack of confocality. This effect is more pronounced in small eyes, such as that of rodents, because the requisite high optical power confers a large dioptric thickness to the retina. A recently-developed structured illumination microscopy (SIM) technique called HiLo imaging has been shown to reduce the effect of out-of-focus light scatter in flood illumination microscopes and produce pseudo-confocal images with significantly improved image quality. In this work, we adopted the HiLo technique to a flood AO ophthalmoscope and performed AO imaging in both (physical) model and live rat eyes. The improvement in image quality from HiLo imaging is shown both qualitatively and quantitatively by using spatial spectral analysis.

  20. Ongoing quality control in digital radiography: Report of AAPM Imaging Physics Committee Task Group 151.

    PubMed

    Jones, A Kyle; Heintz, Philip; Geiser, William; Goldman, Lee; Jerjian, Khachig; Martin, Melissa; Peck, Donald; Pfeiffer, Douglas; Ranger, Nicole; Yorkston, John

    2015-11-01

    Quality control (QC) in medical imaging is an ongoing process and not just a series of infrequent evaluations of medical imaging equipment. The QC process involves designing and implementing a QC program, collecting and analyzing data, investigating results that are outside the acceptance levels for the QC program, and taking corrective action to bring these results back to an acceptable level. The QC process involves key personnel in the imaging department, including the radiologist, radiologic technologist, and the qualified medical physicist (QMP). The QMP performs detailed equipment evaluations and helps with oversight of the QC program, the radiologic technologist is responsible for the day-to-day operation of the QC program. The continued need for ongoing QC in digital radiography has been highlighted in the scientific literature. The charge of this task group was to recommend consistency tests designed to be performed by a medical physicist or a radiologic technologist under the direction of a medical physicist to identify problems with an imaging system that need further evaluation by a medical physicist, including a fault tree to define actions that need to be taken when certain fault conditions are identified. The focus of this final report is the ongoing QC process, including rejected image analysis, exposure analysis, and artifact identification. These QC tasks are vital for the optimal operation of a department performing digital radiography.

  1. Ongoing quality control in digital radiography: Report of AAPM Imaging Physics Committee Task Group 151

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jones, A. Kyle, E-mail: kyle.jones@mdanderson.org; Geiser, William; Heintz, Philip

    Quality control (QC) in medical imaging is an ongoing process and not just a series of infrequent evaluations of medical imaging equipment. The QC process involves designing and implementing a QC program, collecting and analyzing data, investigating results that are outside the acceptance levels for the QC program, and taking corrective action to bring these results back to an acceptable level. The QC process involves key personnel in the imaging department, including the radiologist, radiologic technologist, and the qualified medical physicist (QMP). The QMP performs detailed equipment evaluations and helps with oversight of the QC program, the radiologic technologist ismore » responsible for the day-to-day operation of the QC program. The continued need for ongoing QC in digital radiography has been highlighted in the scientific literature. The charge of this task group was to recommend consistency tests designed to be performed by a medical physicist or a radiologic technologist under the direction of a medical physicist to identify problems with an imaging system that need further evaluation by a medical physicist, including a fault tree to define actions that need to be taken when certain fault conditions are identified. The focus of this final report is the ongoing QC process, including rejected image analysis, exposure analysis, and artifact identification. These QC tasks are vital for the optimal operation of a department performing digital radiography.« less

  2. Learning Receptive Fields and Quality Lookups for Blind Quality Assessment of Stereoscopic Images.

    PubMed

    Shao, Feng; Lin, Weisi; Wang, Shanshan; Jiang, Gangyi; Yu, Mei; Dai, Qionghai

    2016-03-01

    Blind quality assessment of 3D images encounters more new challenges than its 2D counterparts. In this paper, we propose a blind quality assessment for stereoscopic images by learning the characteristics of receptive fields (RFs) from perspective of dictionary learning, and constructing quality lookups to replace human opinion scores without performance loss. The important feature of the proposed method is that we do not need a large set of samples of distorted stereoscopic images and the corresponding human opinion scores to learn a regression model. To be more specific, in the training phase, we learn local RFs (LRFs) and global RFs (GRFs) from the reference and distorted stereoscopic images, respectively, and construct their corresponding local quality lookups (LQLs) and global quality lookups (GQLs). In the testing phase, blind quality pooling can be easily achieved by searching optimal GRF and LRF indexes from the learnt LQLs and GQLs, and the quality score is obtained by combining the LRF and GRF indexes together. Experimental results on three publicly 3D image quality assessment databases demonstrate that in comparison with the existing methods, the devised algorithm achieves high consistent alignment with subjective assessment.

  3. Image Quality in High-resolution and High-cadence Solar Imaging

    NASA Astrophysics Data System (ADS)

    Denker, C.; Dineva, E.; Balthasar, H.; Verma, M.; Kuckein, C.; Diercke, A.; González Manrique, S. J.

    2018-03-01

    Broad-band imaging and even imaging with a moderate bandpass (about 1 nm) provides a photon-rich environment, where frame selection (lucky imaging) becomes a helpful tool in image restoration, allowing us to perform a cost-benefit analysis on how to design observing sequences for imaging with high spatial resolution in combination with real-time correction provided by an adaptive optics (AO) system. This study presents high-cadence (160 Hz) G-band and blue continuum image sequences obtained with the High-resolution Fast Imager (HiFI) at the 1.5-meter GREGOR solar telescope, where the speckle-masking technique is used to restore images with nearly diffraction-limited resolution. The HiFI employs two synchronized large-format and high-cadence sCMOS detectors. The median filter gradient similarity (MFGS) image-quality metric is applied, among others, to AO-corrected image sequences of a pore and a small sunspot observed on 2017 June 4 and 5. A small region of interest, which was selected for fast-imaging performance, covered these contrast-rich features and their neighborhood, which were part of Active Region NOAA 12661. Modifications of the MFGS algorithm uncover the field- and structure-dependency of this image-quality metric. However, MFGS still remains a good choice for determining image quality without a priori knowledge, which is an important characteristic when classifying the huge number of high-resolution images contained in data archives. In addition, this investigation demonstrates that a fast cadence and millisecond exposure times are still insufficient to reach the coherence time of daytime seeing. Nonetheless, the analysis shows that data acquisition rates exceeding 50 Hz are required to capture a substantial fraction of the best seeing moments, significantly boosting the performance of post-facto image restoration.

  4. On pictures and stuff: image quality and material appearance

    NASA Astrophysics Data System (ADS)

    Ferwerda, James A.

    2014-02-01

    Realistic images are a puzzle because they serve as visual representations of objects while also being objects themselves. When we look at an image we are able to perceive both the properties of the image and the properties of the objects represented by the image. Research on image quality has typically focused improving image properties (resolution, dynamic range, frame rate, etc.) while ignoring the issue of whether images are serving their role as visual representations. In this paper we describe a series of experiments that investigate how well images of different quality convey information about the properties of the objects they represent. In the experiments we focus on the effects that two image properties (contrast and sharpness) have on the ability of images to represent the gloss of depicted objects. We found that different experimental methods produced differing results. Specifically, when the stimulus images were presented using simultaneous pair comparison, observers were influenced by the surface properties of the images and conflated changes in image contrast and sharpness with changes in object gloss. On the other hand, when the stimulus images were presented sequentially, observers were able to disregard the image plane properties and more accurately match the gloss of the objects represented by the different quality images. These findings suggest that in understanding image quality it is useful to distinguish between quality of the imaging medium and the quality of the visual information represented by that medium.

  5. Lessions learned in WISE image quality

    NASA Astrophysics Data System (ADS)

    Kendall, Martha; Duval, Valerie G.; Larsen, Mark F.; Heinrichsen, Ingolf H.; Esplin, Roy W.; Shannon, Mark; Wright, Edward L.

    2010-08-01

    The Wide-Field Infrared Survey Explorer (WISE) mission launched in December of 2009 is a true success story. The mission is performing beyond expectations on-orbit and maintained cost and schedule throughout. How does such a thing happen? A team constantly focused on mission success is a key factor. Mission success is more than a program meeting its ultimate science goals; it is also meeting schedule and cost goals to avoid cancellation. The WISE program can attribute some of its success in achieving the image quality needed to meet science goals to lessons learned along the way. A requirement was missed in early decomposition, the absence of which would have adversely affected end-to-end system image quality. Fortunately, the ability of the cross-organizational team to focus on fixing the problem without pointing fingers or waiting for paperwork was crucial in achieving a timely solution. Asking layman questions early in the program could have revealed requirement flowdown misunderstandings between spacecraft control stability and image processing needs. Such is the lesson learned with the WISE spacecraft Attitude Determination & Control Subsystem (ADCS) jitter control and the image data reductions needs. Spacecraft motion can affect image quality in numerous ways. Something as seemingly benign as different terminology being used by teammates in separate groups working on data reduction, spacecraft ADCS, the instrument, mission operations, and the science proved to be a risk to system image quality. While the spacecraft was meeting the allocated jitter requirement , the drift rate variation need was not being met. This missing need was noticed about a year before launch and with a dedicated team effort, an adjustment was made to the spacecraft ADCS control. WISE is meeting all image quality requirements on-orbit thanks to a diligent team noticing something was missing before it was too late and applying their best effort to find a solution.

  6. The Importance of Quality in Ventilation-Perfusion Imaging.

    PubMed

    Mann, April; DiDea, Mario; Fournier, France; Tempesta, Daniel; Williams, Jessica; LaFrance, Norman

    2018-06-01

    As the health care environment continues to change and morph into a system focusing on increased quality and evidence-based outcomes, nuclear medicine technologists must be reminded that they play a critical role in achieving high-quality, interpretable images used to drive patient care, treatment, and best possible outcomes. A survey performed by the Quality Committee of the Society of Nuclear Medicine and Molecular Imaging Technologist Section demonstrated that a clear knowledge gap exists among technologists regarding their understanding of quality, how it is measured, and how it should be achieved by all practicing technologists regardless of role and education level. Understanding of these areas within health care, in conjunction with the growing emphasis on evidence-based outcomes, quality measures, and patient satisfaction, will ultimately elevate the role of nuclear medicine technologists today and into the future. The nuclear medicine role now requires technologists to demonstrate patient assessment skills, practice safety procedures with regard to staff and patients, provide patient education and instruction, and provide physicians with information to assist with the interpretation and outcome of the study. In addition, the technologist must be able to evaluate images by performing technical analysis, knowing the demonstrated anatomy and pathophysiology, and assessing overall quality. Technologists must also be able to triage and understand the disease processes being evaluated and how nuclear medicine diagnostic studies may drive care and treatment. Therefore, it is imperative that nuclear medicine technologists understand their role in the achievement of a high-quality, interpretable study by applying quality principles and understanding and using imaging techniques beyond just basic protocols for every type of disease or system being imaged. This article focuses on quality considerations related to ventilation-perfusion imaging. It provides insight on

  7. Automated image quality assessment for chest CT scans.

    PubMed

    Reeves, Anthony P; Xie, Yiting; Liu, Shuang

    2018-02-01

    Medical image quality needs to be maintained at standards sufficient for effective clinical reading. Automated computer analytic methods may be applied to medical images for quality assessment. For chest CT scans in a lung cancer screening context, an automated quality assessment method is presented that characterizes image noise and image intensity calibration. This is achieved by image measurements in three automatically segmented homogeneous regions of the scan: external air, trachea lumen air, and descending aorta blood. Profiles of CT scanner behavior are also computed. The method has been evaluated on both phantom and real low-dose chest CT scans and results show that repeatable noise and calibration measures may be realized by automated computer algorithms. Noise and calibration profiles show relevant differences between different scanners and protocols. Automated image quality assessment may be useful for quality control for lung cancer screening and may enable performance improvements to automated computer analysis methods. © 2017 American Association of Physicists in Medicine.

  8. Reduced reference image quality assessment via sub-image similarity based redundancy measurement

    NASA Astrophysics Data System (ADS)

    Mou, Xuanqin; Xue, Wufeng; Zhang, Lei

    2012-03-01

    The reduced reference (RR) image quality assessment (IQA) has been attracting much attention from researchers for its loyalty to human perception and flexibility in practice. A promising RR metric should be able to predict the perceptual quality of an image accurately while using as few features as possible. In this paper, a novel RR metric is presented, whose novelty lies in two aspects. Firstly, it measures the image redundancy by calculating the so-called Sub-image Similarity (SIS), and the image quality is measured by comparing the SIS between the reference image and the test image. Secondly, the SIS is computed by the ratios of NSE (Non-shift Edge) between pairs of sub-images. Experiments on two IQA databases (i.e. LIVE and CSIQ databases) show that by using only 6 features, the proposed metric can work very well with high correlations between the subjective and objective scores. In particular, it works consistently well across all the distortion types.

  9. Acceptance testing and commissioning of Kodak Directview CR-850 digital radiography system.

    PubMed

    Bezak, E; Nelligan, R A

    2006-03-01

    This Technical Paper describes Acceptance Testing and Commissioning of the Kodak DirectView CR-850 digital radiography system installed at the Royal Adelaide Hospital. The first of its type installed in Australia, the system is a "dry" image processor, for which no chemicals are required to develop images. Rather, latent radiographic images are stored on photostimulable phosphor screens, which are scanned and displayed by a reader unit. The image can be digitally processed and enhanced before it is forwarded to a storage device, printer or workstation display, thereby alleviating the need to re-expose patients to achieve satisfactory quality images. The phosphor screens are automatically erased, ready for re-use. Results are reported of tests carried out using the optional "Total Quality Tool" quality assurance package installed with the system. This package includes analysis and reporting software which provides for simple testing and reporting of many important characteristics of the system, such as field uniformity, aspect ratio, line and pixel positions, image and system noise, exposure response, scan linearity, modulation transfer function (MTF) and image artefacts. Acceptance Tests were performed for kV and MV exposures. Resolution for MV exposures was at least 0.8 l/mm, and measured phantom dimensions were within 1.05% of expected magnification. Reproducibility between cassettes was within 1.6%. The mean pixel values on the central axis were close to linear for MV exposures from 3 to 10 MU and reached saturation level at around 20 MU for 6 MV and around 30 MV for 23 MV beams. Noise levels were below 0.2 %.

  10. Quality of life and satisfaction with life of malaria patients in context of acceptance of the disease: quantitative studies

    PubMed Central

    2012-01-01

    Background Health status is one of the basic factors of a high quality of life and the problem of the acceptance of illness is important for adaptation to the limitations imposed by it. The purpose of the study was the evaluation of the quality of life, satisfaction with life and the acceptance of illness by malaria patients, as well as the discovery of a relationship between studied parameters. Methods The study was undertaken in August 2010, on 120 Nigerian patients with confirmed malaria. A method of diagnostic survey, based on standardized scales - Acceptance of Illness Scale, The Satisfaction With Life Scale and a standardized survey questionnaire World Health Organization Quality of Life/BREF - was used in this study. Descriptive statistics, variability range, 95% confidence interval, correlation analysis, Spearman’s non-parametric correlation coefficient, Mann–Whitney test and Kruskal-Wallis test were applied and the, so called, test statistics was calculated, followed by the calculation of the test probability p. Results of analyses were presented in a box graph, and a graph of dispersion. Results A dominating share in the adjective scale of the AIS scale was the category of “no acceptance”, given by 71.7% of respondents. The average level of a “somatic domain” was 41.7, and of a “social domain” was 62.8. The mean satisfaction of life evaluation in the SWLS scale was 18 points. The correlation between acceptance of the disease and quality of life for the psychological domain was 0.39***, and between acceptance of the disease and satisfaction with life was 0.40***. The correlation between satisfaction with life and quality of life for the psychological domain was 0.65***, and between satisfaction with life and quality of life for the environment domain was 0.60***. The mean level of AIS for the studied population of men was 16.5, and test probability: p = 0.0014**, and for the environment domain the level was 50, and the test probability

  11. Noise Estimation and Quality Assessment of Gaussian Noise Corrupted Images

    NASA Astrophysics Data System (ADS)

    Kamble, V. M.; Bhurchandi, K.

    2018-03-01

    Evaluating the exact quantity of noise present in an image and quality of an image in the absence of reference image is a challenging task. We propose a near perfect noise estimation method and a no reference image quality assessment method for images corrupted by Gaussian noise. The proposed methods obtain initial estimate of noise standard deviation present in an image using the median of wavelet transform coefficients and then obtains a near to exact estimate using curve fitting. The proposed noise estimation method provides the estimate of noise within average error of +/-4%. For quality assessment, this noise estimate is mapped to fit the Differential Mean Opinion Score (DMOS) using a nonlinear function. The proposed methods require minimum training and yields the noise estimate and image quality score. Images from Laboratory for image and Video Processing (LIVE) database and Computational Perception and Image Quality (CSIQ) database are used for validation of the proposed quality assessment method. Experimental results show that the performance of proposed quality assessment method is at par with the existing no reference image quality assessment metric for Gaussian noise corrupted images.

  12. Technology acceptance and quality of life of the elderly in a telecare program.

    PubMed

    Chou, Chun-Chen; Chang, Chi-Ping; Lee, Ting-Ting; Chou, Hsueh-Fen; Mills, Mary Etta

    2013-07-01

    As information and communication technology applied to telecare has become a trend in elder care services, evaluation of the program is vital for further program design and development. This study evaluated the variables influencing the use of a telecare service program by the elderly. A questionnaire survey was used to explore the relationship between quality of life and technology acceptance of a telecare program in Taiwan. In addition, open-ended questions were used to elicit qualitative information regarding the experience of technology use. The results revealed that elderly persons with better social welfare status and health condition, who used the device frequently, had higher quality of life and accepted technology use. The correlation results also indicated that elderly persons who perceived telecare as useful in solving health problems, had the intention to use the program, and were willing to continue use, also had a better perception toward quality of life regarding their interpersonal relationships and living environment. Nonetheless, cost may be an issue that will impede their willingness to use the technology in care service.

  13. Image quality metrics for volumetric laser displays

    NASA Astrophysics Data System (ADS)

    Williams, Rodney D.; Donohoo, Daniel

    1991-08-01

    This paper addresses the extensions to the image quality metrics and related human factors research that are needed to establish the baseline standards for emerging volume display technologies. The existing and recently developed technologies for multiplanar volume displays are reviewed with an emphasis on basic human visual issues. Human factors image quality metrics and guidelines are needed to firmly establish this technology in the marketplace. The human visual requirements and the display design tradeoffs for these prototype laser-based volume displays are addressed and several critical image quality issues identified for further research. The American National Standard for Human Factors Engineering of Visual Display Terminal Workstations (ANSIHFS-100) and other international standards (ISO, DIN) can serve as a starting point, but this research base must be extended to provide new image quality metrics for this new technology for volume displays.

  14. Image Quality Assessment of High-Resolution Satellite Images with Mtf-Based Fuzzy Comprehensive Evaluation Method

    NASA Astrophysics Data System (ADS)

    Wu, Z.; Luo, Z.; Zhang, Y.; Guo, F.; He, L.

    2018-04-01

    A Modulation Transfer Function (MTF)-based fuzzy comprehensive evaluation method was proposed in this paper for the purpose of evaluating high-resolution satellite image quality. To establish the factor set, two MTF features and seven radiant features were extracted from the knife-edge region of image patch, which included Nyquist, MTF0.5, entropy, peak signal to noise ratio (PSNR), average difference, edge intensity, average gradient, contrast and ground spatial distance (GSD). After analyzing the statistical distribution of above features, a fuzzy evaluation threshold table and fuzzy evaluation membership functions was established. The experiments for comprehensive quality assessment of different natural and artificial objects was done with GF2 image patches. The results showed that the calibration field image has the highest quality scores. The water image has closest image quality to the calibration field, quality of building image is a little poor than water image, but much higher than farmland image. In order to test the influence of different features on quality evaluation, the experiment with different weights were tested on GF2 and SPOT7 images. The results showed that different weights correspond different evaluating effectiveness. In the case of setting up the weights of edge features and GSD, the image quality of GF2 is better than SPOT7. However, when setting MTF and PSNR as main factor, the image quality of SPOT7 is better than GF2.

  15. Cartographic quality of ERTS-1 images

    NASA Technical Reports Server (NTRS)

    Welch, R. I.

    1973-01-01

    Analyses of simulated and operational ERTS images have provided initial estimates of resolution, ground resolution, detectability thresholds and other measures of image quality of interest to earth scientists and cartographers. Based on these values, including an approximate ground resolution of 250 meters for both RBV and MSS systems, the ERTS-1 images appear suited to the production and/or revision of planimetric and photo maps of 1:500,000 scale and smaller for which map accuracy standards are compatible with the imaged detail. Thematic mapping, although less constrained by map accuracy standards, will be influenced by measurement thresholds and errors which have yet to be accurately determined for ERTS images. This study also indicates the desirability of establishing a quantitative relationship between image quality values and map products which will permit both engineers and cartographers/earth scientists to contribute to the design requirements of future satellite imaging systems.

  16. Correlation of radiologists' image quality perception with quantitative assessment parameters: just-noticeable difference vs. peak signal-to-noise ratios

    NASA Astrophysics Data System (ADS)

    Siddiqui, Khan M.; Siegel, Eliot L.; Reiner, Bruce I.; Johnson, Jeffrey P.

    2005-04-01

    The authors identify a fundamental disconnect between the ways in which industry and radiologists assess and even discuss product performance. What is needed is a quantitative methodology that can assess both subjective image quality and observer task performance. In this study, we propose and evaluate the use of a visual discrimination model (VDM) that assesses just-noticeable differences (JNDs) to serve this purpose. The study compares radiologists' subjective perceptions of image quality of computer tomography (CT) and computed radiography (CR) images with quantitative measures of peak signal-to-noise ratio (PSNR) and JNDs as measured by a VDM. The study included 4 CT and 6 CR studies with compression ratios ranging from lossless to 90:1 (total of 80 sets of images were generated [n = 1,200]). Eleven radiologists reviewed the images and rated them in terms of overall quality and readability and identified images not acceptable for interpretation. Normalized reader scores were correlated with compression, objective PSNR, and mean JND values. Results indicated a significantly higher correlation between observer performance and JND values than with PSNR methods. These results support the use of the VDM as a metric not only for the threshold discriminations for which it was calibrated, but also as a general image quality metric. This VDM is a highly promising, reproducible, and reliable adjunct or even alternative to human observer studies for research or to establish clinical guidelines for image compression, dose reductions, and evaluation of various display technologies.

  17. Influence of reconstruction algorithms on image quality in SPECT myocardial perfusion imaging.

    PubMed

    Davidsson, Anette; Olsson, Eva; Engvall, Jan; Gustafsson, Agnetha

    2017-11-01

    We investigated if image- and diagnostic quality in SPECT MPI could be maintained despite a reduced acquisition time adding Depth Dependent Resolution Recovery (DDRR) for image reconstruction. Images were compared with filtered back projection (FBP) and iterative reconstruction using Ordered Subsets Expectation Maximization with (IRAC) and without (IRNC) attenuation correction (AC). Stress- and rest imaging for 15 min was performed on 21 subjects with a dual head gamma camera (Infinia Hawkeye; GE Healthcare), ECG-gating with 8 frames/cardiac cycle and a low-dose CT-scan. A 9 min acquisition was generated using five instead of eight gated frames and was reconstructed with DDRR, with (IRACRR) and without AC (IRNCRR) as well as with FBP. Three experienced nuclear medicine specialists visually assessed anonymized images according to eight criteria on a four point scale, three related to image quality and five to diagnostic confidence. Statistical analysis was performed using Visual Grading Regression (VGR). Observer confidence in statements on image quality was highest for the images that were reconstructed using DDRR (P<0·01 compared to FBP). Iterative reconstruction without DDRR was not superior to FBP. Interobserver variability was significant for statements on image quality (P<0·05) but lower in the diagnostic statements on ischemia and scar. The confidence in assessing ischemia and scar was not different between the reconstruction techniques (P = n.s.). SPECT MPI collected in 9 min, reconstructed with DDRR and AC, produced better image quality than the standard procedure. The observers expressed the highest diagnostic confidence in the DDRR reconstruction. © 2016 Scandinavian Society of Clinical Physiology and Nuclear Medicine. Published by John Wiley & Sons Ltd.

  18. Image gathering and restoration - Information and visual quality

    NASA Technical Reports Server (NTRS)

    Mccormick, Judith A.; Alter-Gartenberg, Rachel; Huck, Friedrich O.

    1989-01-01

    A method is investigated for optimizing the end-to-end performance of image gathering and restoration for visual quality. To achieve this objective, one must inevitably confront the problems that the visual quality of restored images depends on perceptual rather than mathematical considerations and that these considerations vary with the target, the application, and the observer. The method adopted in this paper is to optimize image gathering informationally and to restore images interactively to obtain the visually preferred trade-off among fidelity resolution, sharpness, and clarity. The results demonstrate that this method leads to significant improvements in the visual quality obtained by the traditional digital processing methods. These traditional methods allow a significant loss of visual quality to occur because they treat the design of the image-gathering system and the formulation of the image-restoration algorithm as two separate tasks and fail to account for the transformations between the continuous and the discrete representations in image gathering and reconstruction.

  19. Acceptable distortion and magnification of images on reflective surfaces in an augmented reality system

    NASA Astrophysics Data System (ADS)

    Yamamoto, Shoji; Hosokawa, Natsumi; Yokoya, Mayu; Tsumura, Norimichi

    2016-12-01

    In this paper, we investigated the consistency of visual perception for the change of reflection images in an augmented reality setting. Reflection images with distortion and magnification were generated by changing the capture position of the environment map. Observers evaluated the distortion and magnification in reflection images where the reflected objects were arranged symmetrically or asymmetrically. Our results confirmed that the observers' visual perception was more sensitive to changes in distortion than in magnification in the reflection images. Moreover, the asymmetrical arrangement of reflected objects effectively expands the acceptable range of distortion compared with the symmetrical arrangement.

  20. SU-F-I-46: Optimizing Dose Reduction in Adult Head CT Protocols While Maintaining Image Quality in Postmortem Head Scans

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lipnharski, I; Carranza, C; Quails, N

    Purpose: To optimize adult head CT protocol by reducing dose to an appropriate level while providing CT images of diagnostic quality. Methods: Five cadavers were scanned from the skull base to the vertex using a routine adult head CT protocol (120 kVp, 270 mA, 0.75 s rotation, 0.5 mm × 32 detectors, 70.8 mGy CTDIvol) followed by seven reduced-dose protocols with varying combinations of reduced tube current, reduced rotation time, and increased detectors with CTDIvol ranging from 38.2 to 65.6 mGy. Organ doses were directly measured with 21 OSL dosimeters placed on the surface and implanted in the head bymore » a neurosurgeon. Two neuroradiologists assessed grey-white matter differentiation, fluid space, ventricular size, midline shift, brain mass, edema, ischemia, and skull fractures on a three point scale: (1) Unacceptable, (2) Borderline Acceptable, and (3) Acceptable. Results: For the standard scan, doses to the skin, lens of the eye, salivary glands, thyroid, and brain were 37.55 mGy, 49.65 mGy, 40.67 mGy, 4.63 mGy, and 27.33 mGy, respectively. Two cadavers had cerebral edema due to changing dynamics of postmortem effects, causing the grey-white matter differentiation to appear less distinct. Two cadavers with preserved grey-white matter received acceptable scores for all image quality features for the protocol with a CTDIvol of 57.3 mGy, allowing organ dose savings ranging from 34% to 45%. One cadaver allowed for greater dose reduction for the protocol with a CTDIvol of 42 mGy. Conclusion: Efforts to optimize scan protocol should consider both dose and clinical image quality. This is made possible with postmortem subjects, whose brains are similar to patients, allowing for an investigation of ideal scan parameters. Radiologists at our institution accepted scan protocols acquired with lower scan parameters, with CTDIvol values closer to the American College of Radiology’s (ACR) Achievable Dose level of 57 mGy.« less

  1. PACS: acceptance test, quality control, warranty, and maintenance continuum

    NASA Astrophysics Data System (ADS)

    Romlein, John R.; Norton, Gary S.; Lyche, David K.; Richardson, Ronald R., Jr.

    1999-07-01

    As PACS gain greater acceptance and use in medical facilities the question of life cycle management must be addressed in terms that relate to the common business practices for medical information system and medical devices. The issues in life cycle management of such a system are relatively new to the industry. Increased use of PACS within the medical community requires that standardized life cycle management practices by developed and implemented. This paper develops a new of life cycle issues as cyclic and related events that are not only manageable, but also predictable in terms, of, frequency, duration data content, data exchange, potential outcomes, staffing requirement, documentation, and staff interaction. This view is presented as a continuum that begins at the acceptance testing of a PACS and continues throughout its life cycle. The continuum incorporates the required relationship between quality control testing and maintenance actions during warranty period and the maintenance years. Interrelated cyclic events are described that bind these processes together and provide a basis for long-term proactive management of PACS in a medical environment.

  2. Ultra-low-dose lung screening CT with model-based iterative reconstruction: an assessment of image quality and lesion conspicuity.

    PubMed

    Ju, Yun Hye; Lee, Geewon; Lee, Ji Won; Hong, Seung Baek; Suh, Young Ju; Jeong, Yeon Joo

    2018-05-01

    Background Reducing radiation dose inevitably increases image noise, and thus, it is important in low-dose computed tomography (CT) to maintain image quality and lesion detection performance. Purpose To assess image quality and lesion conspicuity of ultra-low-dose CT with model-based iterative reconstruction (MBIR) and to determine a suitable protocol for lung screening CT. Material and Methods A total of 120 heavy smokers underwent lung screening CT and were randomly and equally assigned to one of five groups: group 1 = 120 kVp, 25 mAs, with FBP reconstruction; group 2 = 120 kVp, 10 mAs, with MBIR; group 3 = 100 kVp, 15 mAs, with MBIR; group 4 = 100 kVp, 10 mAs, with MBIR; and group 5 = 100 kVp, 5 mAs, with MBIR. Two radiologists evaluated intergroup differences with respect to radiation dose, image noise, image quality, and lesion conspicuity using the Kruskal-Wallis test and the Chi-square test. Results Effective doses were 61-87% lower in groups 2-5 than in group 1. Image noises in groups 1 and 5 were significantly higher than in the other groups ( P < 0.001). Overall image quality was best in group 1, but diagnostic acceptability of overall image qualities in groups 1-3 was not significantly different (all P values > 0.05). Lesion conspicuities were similar in groups 1-4, but were significantly poorer in group 5. Conclusion Lung screening CT with MBIR obtained at 100 kVp and 15 mAs enables a ∼60% reduction in radiation dose versus low-dose CT, while maintaining image quality and lesion conspicuity.

  3. The effect of image quality and forensic expertise in facial image comparisons.

    PubMed

    Norell, Kristin; Läthén, Klas Brorsson; Bergström, Peter; Rice, Allyson; Natu, Vaidehi; O'Toole, Alice

    2015-03-01

    Images of perpetrators in surveillance video footage are often used as evidence in court. In this study, identification accuracy was compared for forensic experts and untrained persons in facial image comparisons as well as the impact of image quality. Participants viewed thirty image pairs and were asked to rate the level of support garnered from their observations for concluding whether or not the two images showed the same person. Forensic experts reached their conclusions with significantly fewer errors than did untrained participants. They were also better than novices at determining when two high-quality images depicted the same person. Notably, lower image quality led to more careful conclusions by experts, but not for untrained participants. In summary, the untrained participants had more false negatives and false positives than experts, which in the latter case could lead to a higher risk of an innocent person being convicted for an untrained witness. © 2014 American Academy of Forensic Sciences.

  4. High-volume image quality assessment systems: tuning performance with an interactive data visualization tool

    NASA Astrophysics Data System (ADS)

    Bresnahan, Patricia A.; Pukinskis, Madeleine; Wiggins, Michael

    1999-03-01

    Image quality assessment systems differ greatly with respect to the number and types of mags they need to evaluate, and their overall architectures. Managers of these systems, however, all need to be able to tune and evaluate system performance, requirements often overlooked or under-designed during project planning. Performance tuning tools allow users to define acceptable quality standards for image features and attributes by adjusting parameter settings. Performance analysis tools allow users to evaluate and/or predict how well a system performs in a given parameter state. While image assessment algorithms are becoming quite sophisticated, duplicating or surpassing the human decision making process in their speed and reliability, they often require a greater investment in 'training' or fine tuning of parameters in order to achieve optimum performance. This process may involve the analysis of hundreds or thousands of images, generating a large database of files and statistics that can be difficult to sort through and interpret. Compounding the difficulty is the fact that personnel charged with tuning and maintaining the production system may not have the statistical or analytical background required for the task. Meanwhile, hardware innovations have greatly increased the volume of images that can be handled in a given time frame, magnifying the consequences of running a production site with an inadequately tuned system. In this paper, some general requirements for a performance evaluation and tuning data visualization system are discussed. A custom engineered solution to the tuning and evaluation problem is then presented, developed within the context of a high volume image quality assessment, data entry, OCR, and image archival system. A key factor influencing the design of the system was the context-dependent definition of image quality, as perceived by a human interpreter. This led to the development of a five-level, hierarchical approach to image quality

  5. Blind image quality assessment without training on human opinion scores

    NASA Astrophysics Data System (ADS)

    Mittal, Anish; Soundararajan, Rajiv; Muralidhar, Gautam S.; Bovik, Alan C.; Ghosh, Joydeep

    2013-03-01

    We propose a family of image quality assessment (IQA) models based on natural scene statistics (NSS), that can predict the subjective quality of a distorted image without reference to a corresponding distortionless image, and without any training results on human opinion scores of distorted images. These `completely blind' models compete well with standard non-blind image quality indices in terms of subjective predictive performance when tested on the large publicly available `LIVE' Image Quality database.

  6. Quality measures in applications of image restoration.

    PubMed

    Kriete, A; Naim, M; Schafer, L

    2001-01-01

    We describe a new method for the estimation of image quality in image restoration applications. We demonstrate this technique on a simulated data set of fluorescent beads, in comparison with restoration by three different deconvolution methods. Both the number of iterations and a regularisation factor are varied to enforce changes in the resulting image quality. First, the data sets are directly compared by an accuracy measure. These values serve to validate the image quality descriptor, which is developed on the basis of optical information theory. This most general measure takes into account the spectral energies and the noise, weighted in a logarithmic fashion. It is demonstrated that this method is particularly helpful as a user-oriented method to control the output of iterative image restorations and to eliminate the guesswork in choosing a suitable number of iterations.

  7. Image quality, threshold contrast and mean glandular dose in CR mammography

    NASA Astrophysics Data System (ADS)

    Jakubiak, R. R.; Gamba, H. R.; Neves, E. B.; Peixoto, J. E.

    2013-09-01

    In many countries, computed radiography (CR) systems represent the majority of equipment used in digital mammography. This study presents a method for optimizing image quality and dose in CR mammography of patients with breast thicknesses between 45 and 75 mm. Initially, clinical images of 67 patients (group 1) were analyzed by three experienced radiologists, reporting about anatomical structures, noise and contrast in low and high pixel value areas, and image sharpness and contrast. Exposure parameters (kV, mAs and target/filter combination) used in the examinations of these patients were reproduced to determine the contrast-to-noise ratio (CNR) and mean glandular dose (MGD). The parameters were also used to radiograph a CDMAM (version 3.4) phantom (Artinis Medical Systems, The Netherlands) for image threshold contrast evaluation. After that, different breast thicknesses were simulated with polymethylmethacrylate layers and various sets of exposure parameters were used in order to determine optimal radiographic parameters. For each simulated breast thickness, optimal beam quality was defined as giving a target CNR to reach the threshold contrast of CDMAM images for acceptable MGD. These results were used for adjustments in the automatic exposure control (AEC) by the maintenance team. Using optimized exposure parameters, clinical images of 63 patients (group 2) were evaluated as described above. Threshold contrast, CNR and MGD for such exposure parameters were also determined. Results showed that the proposed optimization method was effective for all breast thicknesses studied in phantoms. The best result was found for breasts of 75 mm. While in group 1 there was no detection of the 0.1 mm critical diameter detail with threshold contrast below 23%, after the optimization, detection occurred in 47.6% of the images. There was also an average MGD reduction of 7.5%. The clinical image quality criteria were attended in 91.7% for all breast thicknesses evaluated in both

  8. Image quality classification for DR screening using deep learning.

    PubMed

    FengLi Yu; Jing Sun; Annan Li; Jun Cheng; Cheng Wan; Jiang Liu

    2017-07-01

    The quality of input images significantly affects the outcome of automated diabetic retinopathy (DR) screening systems. Unlike the previous methods that only consider simple low-level features such as hand-crafted geometric and structural features, in this paper we propose a novel method for retinal image quality classification (IQC) that performs computational algorithms imitating the working of the human visual system. The proposed algorithm combines unsupervised features from saliency map and supervised features coming from convolutional neural networks (CNN), which are fed to an SVM to automatically detect high quality vs poor quality retinal fundus images. We demonstrate the superior performance of our proposed algorithm on a large retinal fundus image dataset and the method could achieve higher accuracy than other methods. Although retinal images are used in this study, the methodology is applicable to the image quality assessment and enhancement of other types of medical images.

  9. Staff acceptance of video monitoring for coordination: a video system to support perioperative situation awareness.

    PubMed

    Kim, Young Ju; Xiao, Yan; Hu, Peter; Dutton, Richard

    2009-08-01

    To understand staff acceptance of a remote video monitoring system for operating room (OR) coordination. Improved real-time remote visual access to OR may enhance situational awareness but also raises privacy concerns for patients and staff. Survey. A system was implemented in a six-room surgical suite to display OR monitoring video at an access restricted control desk area. Image quality was manipulated to improve staff acceptance. Two months after installation, interviews and a survey were conducted on staff acceptance of video monitoring. About half of all OR personnel responded (n = 63). Overall levels of concerns were low, with 53% rated no concerns and 42% little concern. Top two reported uses of the video were to see if cases are finished and to see if a room is ready. Viewing the video monitoring system as useful did not reduce levels of concern. Staff in supervisory positions perceived less concern about the system's impact on privacy than did those supervised (p < 0.03). Concerns for patient privacy correlated with concerns for staff privacy and performance monitoring. Technical means such as manipulating image quality helped staff acceptance. Manipulation of image quality resulted overall acceptance of monitoring video, with residual levels of concerns. OR nurses may express staff privacy concern in the form of concerns over patient privacy. This study provided suggestions for technological and implementation strategies of video monitoring for coordination use in OR. Deployment of communication technology and integration of clinical information will likely raise concerns over staff privacy and performance monitoring. The potential gain of increased information access may be offset by negative impact of a sense of loss of autonomy.

  10. The relationship between physician burnout and quality of healthcare in terms of safety and acceptability: a systematic review

    PubMed Central

    Dewa, Carolyn S; Loong, Desmond; Bonato, Sarah; Trojanowski, Lucy

    2017-01-01

    Objectives This study reviews the current state of the published peer-reviewed literature related to physician burnout and two quality of care dimensions. The purpose of this systematic literature review is to address the question, ‘How does physician burnout affect the quality of healthcare related to the dimensions of acceptability and safety?’ Design Using a multiphase screening process, this systematic literature review is based on publically available peer-reviewed studies published between 2002 and 2017. Six electronic databases were searched: (1) MEDLINE Current, (2) MEDLINE In-process, (3) MEDLINE Epub Ahead of Print, (4) PsycINFO, (5) Embase and (6) Web of Science. Setting Physicians practicing in civilian settings. Participants Practicing physicians who have completed training. Primary and secondary outcome measures Quality of healthcare related to acceptability (ie, patient satisfaction, physician communication and physician attitudes) and safety (ie, minimising risks or harm to patients). Results 4114 unique citations were identified. Of these, 12 articles were included in the review. Two studies were rated as having high risk of bias and 10 as having moderate risk. Four studies were conducted in North America, four in Europe, one in the Middle East and three in East Asia. Results of this systematic literature review suggest there is moderate evidence that burnout is associated with safety-related quality of care. Because of the variability in the way patient acceptability-related quality of care was measured and the inconsistency in study findings, the evidence supporting the relationship between burnout and patient acceptability-related quality of care is less strong. Conclusions The focus on direct care-related quality highlights additional ways that physician burnout affects the healthcare system. These studies can help to inform decisions about how to improve patient care by addressing physician burnout. Continued work looking at the

  11. Image quality assessment and medical physics evaluation of different portable dental X-ray units.

    PubMed

    Pittayapat, Pisha; Oliveira-Santos, Christiano; Thevissen, Patrick; Michielsen, Koen; Bergans, Niki; Willems, Guy; Debruyckere, Deborah; Jacobs, Reinhilde

    2010-09-10

    using remote control (distance>1m: Rextar <0.2 microGy, MinRay <0.1 microGy). The present study demonstrated the feasibility of three portable X-ray systems to be used for specific indications, based on acceptable image quality and sufficient accuracy of the machines and following the standard guidelines for radiation hygiene. Copyright 2010 Elsevier Ireland Ltd. All rights reserved.

  12. Comparison of the performance of intraoral X-ray sensors using objective image quality assessment.

    PubMed

    Hellén-Halme, Kristina; Johansson, Curt; Nilsson, Mats

    2016-05-01

    The main aim of this study was to evaluate the performance of 10 individual sensors of the same make, using objective measures of key image quality parameters. A further aim was to compare 8 brands of sensors. Ten new sensors of 8 different models from 6 manufacturers (i.e., 80 sensors) were included in the study. All sensors were exposed in a standardized way using an X-ray tube voltage of 60 kVp and different exposure times. Sensor response, noise, low-contrast resolution, spatial resolution and uniformity were measured. Individual differences between sensors of the same brand were surprisingly large in some cases. There were clear differences in the characteristics of the different brands of sensors. The largest variations were found for individual sensor response for some of the brands studied. Also, noise level and low contrast resolution showed large variations between brands. Sensors, even of the same brand, vary significantly in their quality. It is thus valuable to establish action levels for the acceptance of newly delivered sensors and to use objective image quality control for commissioning purposes and periodic checks to ensure high performance of individual digital sensors. Copyright © 2016 Elsevier Inc. All rights reserved.

  13. Image aesthetic quality evaluation using convolution neural network embedded learning

    NASA Astrophysics Data System (ADS)

    Li, Yu-xin; Pu, Yuan-yuan; Xu, Dan; Qian, Wen-hua; Wang, Li-peng

    2017-11-01

    A way of embedded learning convolution neural network (ELCNN) based on the image content is proposed to evaluate the image aesthetic quality in this paper. Our approach can not only solve the problem of small-scale data but also score the image aesthetic quality. First, we chose Alexnet and VGG_S to compare for confirming which is more suitable for this image aesthetic quality evaluation task. Second, to further boost the image aesthetic quality classification performance, we employ the image content to train aesthetic quality classification models. But the training samples become smaller and only using once fine-tuning cannot make full use of the small-scale data set. Third, to solve the problem in second step, a way of using twice fine-tuning continually based on the aesthetic quality label and content label respective is proposed, the classification probability of the trained CNN models is used to evaluate the image aesthetic quality. The experiments are carried on the small-scale data set of Photo Quality. The experiment results show that the classification accuracy rates of our approach are higher than the existing image aesthetic quality evaluation approaches.

  14. The Detection of Focal Liver Lesions Using Abdominal CT: A Comparison of Image Quality Between Adaptive Statistical Iterative Reconstruction V and Adaptive Statistical Iterative Reconstruction.

    PubMed

    Lee, Sangyun; Kwon, Heejin; Cho, Jihan

    2016-12-01

    To investigate image quality characteristics of abdominal computed tomography (CT) scans reconstructed with adaptive statistical iterative reconstruction V (ASIR-V) vs currently using applied adaptive statistical iterative reconstruction (ASIR). This institutional review board-approved study included 35 consecutive patients who underwent CT of the abdomen. Among these 35 patients, 27 with focal liver lesions underwent abdomen CT with a 128-slice multidetector unit using the following parameters: fixed noise index of 30, 1.25 mm slice thickness, 120 kVp, and a gantry rotation time of 0.5 seconds. CT images were analyzed depending on the method of reconstruction: ASIR (30%, 50%, and 70%) vs ASIR-V (30%, 50%, and 70%). Three radiologists independently assessed randomized images in a blinded manner. Imaging sets were compared to focal lesion detection numbers, overall image quality, and objective noise with a paired sample t test. Interobserver agreement was assessed with the intraclass correlation coefficient. The detection of small focal liver lesions (<10 mm) was significantly higher when ASIR-V was used when compared to ASIR (P <0.001). Subjective image noise, artifact, and objective image noise in liver were generally significantly better for ASIR-V compared to ASIR, especially in 50% ASIR-V. Image sharpness and diagnostic acceptability were significantly worse in 70% ASIR-V compared to various levels of ASIR. Images analyzed using 50% ASIR-V were significantly better than three different series of ASIR or other ASIR-V conditions at providing diagnostically acceptable CT scans without compromising image quality and in the detection of focal liver lesions. Copyright © 2016 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

  15. Cross-layer Energy Optimization Under Image Quality Constraints for Wireless Image Transmissions.

    PubMed

    Yang, Na; Demirkol, Ilker; Heinzelman, Wendi

    2012-01-01

    Wireless image transmission is critical in many applications, such as surveillance and environment monitoring. In order to make the best use of the limited energy of the battery-operated cameras, while satisfying the application-level image quality constraints, cross-layer design is critical. In this paper, we develop an image transmission model that allows the application layer (e.g., the user) to specify an image quality constraint, and optimizes the lower layer parameters of transmit power and packet length, to minimize the energy dissipation in image transmission over a given distance. The effectiveness of this approach is evaluated by applying the proposed energy optimization to a reference ZigBee system and a WiFi system, and also by comparing to an energy optimization study that does not consider any image quality constraint. Evaluations show that our scheme outperforms the default settings of the investigated commercial devices and saves a significant amount of energy at middle-to-large transmission distances.

  16. Retinal Image Quality Assessment for Spaceflight-Induced Vision Impairment Study

    NASA Technical Reports Server (NTRS)

    Vu, Amanda Cadao; Raghunandan, Sneha; Vyas, Ruchi; Radhakrishnan, Krishnan; Taibbi, Giovanni; Vizzeri, Gianmarco; Grant, Maria; Chalam, Kakarla; Parsons-Wingerter, Patricia

    2015-01-01

    Long-term exposure to space microgravity poses significant risks for visual impairment. Evidence suggests such vision changes are linked to cephalad fluid shifts, prompting a need to directly quantify microgravity-induced retinal vascular changes. The quality of retinal images used for such vascular remodeling analysis, however, is dependent on imaging methodology. For our exploratory study, we hypothesized that retinal images captured using fluorescein imaging methodologies would be of higher quality in comparison to images captured without fluorescein. A semi-automated image quality assessment was developed using Vessel Generation Analysis (VESGEN) software and MATLAB® image analysis toolboxes. An analysis of ten images found that the fluorescein imaging modality provided a 36% increase in overall image quality (two-tailed p=0.089) in comparison to nonfluorescein imaging techniques.

  17. Multi-institutional MicroCT image comparison of image-guided small animal irradiators

    NASA Astrophysics Data System (ADS)

    Johnstone, Chris D.; Lindsay, Patricia; E Graves, Edward; Wong, Eugene; Perez, Jessica R.; Poirier, Yannick; Ben-Bouchta, Youssef; Kanesalingam, Thilakshan; Chen, Haijian; E Rubinstein, Ashley; Sheng, Ke; Bazalova-Carter, Magdalena

    2017-07-01

    To recommend imaging protocols and establish tolerance levels for microCT image quality assurance (QA) performed on conformal image-guided small animal irradiators. A fully automated QA software SAPA (small animal phantom analyzer) for image analysis of the commercial Shelley micro-CT MCTP 610 phantom was developed, in which quantitative analyses of CT number linearity, signal-to-noise ratio (SNR), uniformity and noise, geometric accuracy, spatial resolution by means of modulation transfer function (MTF), and CT contrast were performed. Phantom microCT scans from eleven institutions acquired with four image-guided small animal irradiator units (including the commercial PXi X-RAD SmART and Xstrahl SARRP systems) with varying parameters used for routine small animal imaging were analyzed. Multi-institutional data sets were compared using SAPA, based on which tolerance levels for each QA test were established and imaging protocols for QA were recommended. By analyzing microCT data from 11 institutions, we established image QA tolerance levels for all image quality tests. CT number linearity set to R 2  >  0.990 was acceptable in microCT data acquired at all but three institutions. Acceptable SNR  >  36 and noise levels  <55 HU were obtained at five of the eleven institutions, where failing scans were acquired with current-exposure time of less than 120 mAs. Acceptable spatial resolution (>1.5 lp mm-1 for MTF  =  0.2) was obtained at all but four institutions due to their large image voxel size used (>0.275 mm). Ten of the eleven institutions passed the set QA tolerance for geometric accuracy (<1.5%) and nine of the eleven institutions passed the QA tolerance for contrast (>2000 HU for 30 mgI ml-1). We recommend performing imaging QA with 70 kVp, 1.5 mA, 120 s imaging time, 0.20 mm voxel size, and a frame rate of 5 fps for the PXi X-RAD SmART. For the Xstrahl SARRP, we recommend using 60 kVp, 1.0 mA, 240 s imaging time, 0.20

  18. Quality Improvement of Liver Ultrasound Images Using Fuzzy Techniques.

    PubMed

    Bayani, Azadeh; Langarizadeh, Mostafa; Radmard, Amir Reza; Nejad, Ahmadreza Farzaneh

    2016-12-01

    Liver ultrasound images are so common and are applied so often to diagnose diffuse liver diseases like fatty liver. However, the low quality of such images makes it difficult to analyze them and diagnose diseases. The purpose of this study, therefore, is to improve the contrast and quality of liver ultrasound images. In this study, a number of image contrast enhancement algorithms which are based on fuzzy logic were applied to liver ultrasound images - in which the view of kidney is observable - using Matlab2013b to improve the image contrast and quality which has a fuzzy definition; just like image contrast improvement algorithms using a fuzzy intensification operator, contrast improvement algorithms applying fuzzy image histogram hyperbolization, and contrast improvement algorithms by fuzzy IF-THEN rules. With the measurement of Mean Squared Error and Peak Signal to Noise Ratio obtained from different images, fuzzy methods provided better results, and their implementation - compared with histogram equalization method - led both to the improvement of contrast and visual quality of images and to the improvement of liver segmentation algorithms results in images. Comparison of the four algorithms revealed the power of fuzzy logic in improving image contrast compared with traditional image processing algorithms. Moreover, contrast improvement algorithm based on a fuzzy intensification operator was selected as the strongest algorithm considering the measured indicators. This method can also be used in future studies on other ultrasound images for quality improvement and other image processing and analysis applications.

  19. Designing a two-rank acceptance sampling plan for quality inspection of geospatial data products

    NASA Astrophysics Data System (ADS)

    Tong, Xiaohua; Wang, Zhenhua; Xie, Huan; Liang, Dan; Jiang, Zuoqin; Li, Jinchao; Li, Jun

    2011-10-01

    To address the disadvantages of classical sampling plans designed for traditional industrial products, we originally propose a two-rank acceptance sampling plan (TRASP) for the inspection of geospatial data outputs based on the acceptance quality level (AQL). The first rank sampling plan is to inspect the lot consisting of map sheets, and the second is to inspect the lot consisting of features in an individual map sheet. The TRASP design is formulated as an optimization problem with respect to sample size and acceptance number, which covers two lot size cases. The first case is for a small lot size with nonconformities being modeled by a hypergeometric distribution function, and the second is for a larger lot size with nonconformities being modeled by a Poisson distribution function. The proposed TRASP is illustrated through two empirical case studies. Our analysis demonstrates that: (1) the proposed TRASP provides a general approach for quality inspection of geospatial data outputs consisting of non-uniform items and (2) the proposed acceptance sampling plan based on TRASP performs better than other classical sampling plans. It overcomes the drawbacks of percent sampling, i.e., "strictness for large lot size, toleration for small lot size," and those of a national standard used specifically for industrial outputs, i.e., "lots with different sizes corresponding to the same sampling plan."

  20. Intimate relationship quality, self-concept and illness acceptance in those with multiple sclerosis.

    PubMed

    Wright, Thomas M; Kiropoulos, Litza A

    2017-02-01

    Lower levels of Intimate Relationship Quality (IRQ) have been found in those with Multiple Sclerosis (MS) compared to the general population. This study examined an MS sample to see whether IRQ was positively associated with self-concept, whether IRQ was positively associated with MS illness acceptance and whether IRQ was predicted by self-concept and illness acceptance. In this cross-sectional study, 115 participants with MS who were in an intimate relationship completed an online survey advertised on MS related websites. The survey assessed demographic variables, MS illness variables and levels of IRQ, self-concept and illness acceptance. Results revealed that IRQ was significantly positively associated with self-concept and with illness acceptance. Multiple hierarchical linear regression analysis revealed that, after controlling for illness duration and level of disability, self-concept significantly predicted IRQ but illness acceptance did not significantly predict IRQ. This study addressed several gaps and methodological flaws in the literature and was the first known to assess predictors of IRQ in those with MS. The results suggest that self-concept could be a potential target for individual and couple psychological interventions to improve IRQ and contribute to improved outcomes for those with MS.

  1. Quality evaluation of no-reference MR images using multidirectional filters and image statistics.

    PubMed

    Jang, Jinseong; Bang, Kihun; Jang, Hanbyol; Hwang, Dosik

    2018-09-01

    This study aimed to develop a fully automatic, no-reference image-quality assessment (IQA) method for MR images. New quality-aware features were obtained by applying multidirectional filters to MR images and examining the feature statistics. A histogram of these features was then fitted to a generalized Gaussian distribution function for which the shape parameters yielded different values depending on the type of distortion in the MR image. Standard feature statistics were established through a training process based on high-quality MR images without distortion. Subsequently, the feature statistics of a test MR image were calculated and compared with the standards. The quality score was calculated as the difference between the shape parameters of the test image and the undistorted standard images. The proposed IQA method showed a >0.99 correlation with the conventional full-reference assessment methods; accordingly, this proposed method yielded the best performance among no-reference IQA methods for images containing six types of synthetic, MR-specific distortions. In addition, for authentically distorted images, the proposed method yielded the highest correlation with subjective assessments by human observers, thus demonstrating its superior performance over other no-reference IQAs. Our proposed IQA was designed to consider MR-specific features and outperformed other no-reference IQAs designed mainly for photographic images. Magn Reson Med 80:914-924, 2018. © 2018 International Society for Magnetic Resonance in Medicine. © 2018 International Society for Magnetic Resonance in Medicine.

  2. Validation of no-reference image quality index for the assessment of digital mammographic images

    NASA Astrophysics Data System (ADS)

    de Oliveira, Helder C. R.; Barufaldi, Bruno; Borges, Lucas R.; Gabarda, Salvador; Bakic, Predrag R.; Maidment, Andrew D. A.; Schiabel, Homero; Vieira, Marcelo A. C.

    2016-03-01

    To ensure optimal clinical performance of digital mammography, it is necessary to obtain images with high spatial resolution and low noise, keeping radiation exposure as low as possible. These requirements directly affect the interpretation of radiologists. The quality of a digital image should be assessed using objective measurements. In general, these methods measure the similarity between a degraded image and an ideal image without degradation (ground-truth), used as a reference. These methods are called Full-Reference Image Quality Assessment (FR-IQA). However, for digital mammography, an image without degradation is not available in clinical practice; thus, an objective method to assess the quality of mammograms must be performed without reference. The purpose of this study is to present a Normalized Anisotropic Quality Index (NAQI), based on the Rényi entropy in the pseudo-Wigner domain, to assess mammography images in terms of spatial resolution and noise without any reference. The method was validated using synthetic images acquired through an anthropomorphic breast software phantom, and the clinical exposures on anthropomorphic breast physical phantoms and patient's mammograms. The results reported by this noreference index follow the same behavior as other well-established full-reference metrics, e.g., the peak signal-to-noise ratio (PSNR) and structural similarity index (SSIM). Reductions of 50% on the radiation dose in phantom images were translated as a decrease of 4dB on the PSNR, 25% on the SSIM and 33% on the NAQI, evidencing that the proposed metric is sensitive to the noise resulted from dose reduction. The clinical results showed that images reduced to 53% and 30% of the standard radiation dose reported reductions of 15% and 25% on the NAQI, respectively. Thus, this index may be used in clinical practice as an image quality indicator to improve the quality assurance programs in mammography; hence, the proposed method reduces the subjectivity

  3. Quality Improvement of Liver Ultrasound Images Using Fuzzy Techniques

    PubMed Central

    Bayani, Azadeh; Langarizadeh, Mostafa; Radmard, Amir Reza; Nejad, Ahmadreza Farzaneh

    2016-01-01

    Background: Liver ultrasound images are so common and are applied so often to diagnose diffuse liver diseases like fatty liver. However, the low quality of such images makes it difficult to analyze them and diagnose diseases. The purpose of this study, therefore, is to improve the contrast and quality of liver ultrasound images. Methods: In this study, a number of image contrast enhancement algorithms which are based on fuzzy logic were applied to liver ultrasound images - in which the view of kidney is observable - using Matlab2013b to improve the image contrast and quality which has a fuzzy definition; just like image contrast improvement algorithms using a fuzzy intensification operator, contrast improvement algorithms applying fuzzy image histogram hyperbolization, and contrast improvement algorithms by fuzzy IF-THEN rules. Results: With the measurement of Mean Squared Error and Peak Signal to Noise Ratio obtained from different images, fuzzy methods provided better results, and their implementation - compared with histogram equalization method - led both to the improvement of contrast and visual quality of images and to the improvement of liver segmentation algorithms results in images. Conclusion: Comparison of the four algorithms revealed the power of fuzzy logic in improving image contrast compared with traditional image processing algorithms. Moreover, contrast improvement algorithm based on a fuzzy intensification operator was selected as the strongest algorithm considering the measured indicators. This method can also be used in future studies on other ultrasound images for quality improvement and other image processing and analysis applications. PMID:28077898

  4. Dose and diagnostic image quality in digital tomosynthesis imaging of facial bones in pediatrics

    NASA Astrophysics Data System (ADS)

    King, J. M.; Hickling, S.; Elbakri, I. A.; Reed, M.; Wrogemann, J.

    2011-03-01

    The purpose of this study was to evaluate the use of digital tomosynthesis (DT) for pediatric facial bone imaging. We compared the eye lens dose and diagnostic image quality of DT facial bone exams relative to digital radiography (DR) and computed tomography (CT), and investigated whether we could modify our current DT imaging protocol to reduce patient dose while maintaining sufficient diagnostic image quality. We measured the dose to the eye lens for all three modalities using high-sensitivity thermoluminescent dosimeters (TLDs) and an anthropomorphic skull phantom. To assess the diagnostic image quality of DT compared to the corresponding DR and CT images, we performed an observer study where the visibility of anatomical structures in the DT phantom images were rated on a four-point scale. We then acquired DT images at lower doses and had radiologists indicate whether the visibility of each structure was adequate for diagnostic purposes. For typical facial bone exams, we measured eye lens doses of 0.1-0.4 mGy for DR, 0.3-3.7 mGy for DT, and 26 mGy for CT. In general, facial bone structures were visualized better with DT then DR, and the majority of structures were visualized well enough to avoid the need for CT. DT imaging provides high quality diagnostic images of the facial bones while delivering significantly lower doses to the lens of the eye compared to CT. In addition, we found that by adjusting the imaging parameters, the DT effective dose can be reduced by up to 50% while maintaining sufficient image quality.

  5. Imaging quality evaluation method of pixel coupled electro-optical imaging system

    NASA Astrophysics Data System (ADS)

    He, Xu; Yuan, Li; Jin, Chunqi; Zhang, Xiaohui

    2017-09-01

    With advancements in high-resolution imaging optical fiber bundle fabrication technology, traditional photoelectric imaging system have become ;flexible; with greatly reduced volume and weight. However, traditional image quality evaluation models are limited by the coupling discrete sampling effect of fiber-optic image bundles and charge-coupled device (CCD) pixels. This limitation substantially complicates the design, optimization, assembly, and evaluation image quality of the coupled discrete sampling imaging system. Based on the transfer process of grayscale cosine distribution optical signal in the fiber-optic image bundle and CCD, a mathematical model of coupled modulation transfer function (coupled-MTF) is established. This model can be used as a basis for following studies on the convergence and periodically oscillating characteristics of the function. We also propose the concept of the average coupled-MTF, which is consistent with the definition of traditional MTF. Based on this concept, the relationships among core distance, core layer radius, and average coupled-MTF are investigated.

  6. Patients after colostomy: relationship between quality of life and acceptance of disability and social support.

    PubMed

    Zhang, Tie-Ling; Hu, Ai-Ling; Xu, Hong-Lian; Zheng, Mei-Chun; Liang, Ming-Juan

    2013-11-01

    The aim of this research was to explore quality of life (QOL) and acceptance of disability and social support of colostomy patients as well as the relationship between these factors. A descriptive, correlational study was conducted using four scales: the European Organization for Research and Treatment of Cancer Quality of Life Questionnaire (EORTC QLQ-C30) and European Organization for Research and Treatment of Cancer Colorectal Cancer-Specific Quality of Life Questionnaire (EORTC QLQ-CR38) scales, the Acceptance of Disability Scale (ADS), and the Social Relational Quality Scale (SRQS). A convenience sample of 111 colostomy patients from four hospitals in Guangzhou who underwent colostomy operation at least one month prior to the study and who visited the stoma clinic or association from August 2011 to February 2012 was evaluated for inclusion in the study. All statistical analyses were performed using SPSS 17.0 software (SPSS Inc., Chicago, IL, USA). The patients' general health status was better than the reference level recommended by European Organization for Research and Treatment of Cancer, and the overall ADS score was average. The SRQS score was similar to that found in a Hong Kong study. The general health status and dimensions of QOL were significantly correlated with ADS and all of its dimensions (P < 0.05). The general health status and dimensions of QOL were also significantly correlated with SRQS and all of its dimensions (P < 0.05). QOL, acceptance of disability, and social relational quality of colostomy patients were closely related. Our results emphasize that patients should work to form rational values and close bonds with families and friends to achieve a better QOL.

  7. [Quality control of laser imagers].

    PubMed

    Winkelbauer, F; Ammann, M; Gerstner, N; Imhof, H

    1992-11-01

    Multiformat imagers based on laser systems are used for documentation in an increasing number of investigations. The specific problems of quality control are explained and the persistence of film processing in these imager systems of different configuration with (Machine 1: 3M-Laser-Imager-Plus M952 with connected 3M Film-Processor, 3M-Film IRB, X-Rax Chemical Mixer 3M-XPM, 3M-Developer and Fixer) or without (Machine 2: 3M-Laser-Imager-Plus M952 with separate DuPont-Cronex Film-processor, Kodak IR-Film, Kodak Automixer, Kodak-Developer and Fixer) connected film processing unit are investigated. In our checking based on DIN 6868 and ONORM S 5240 we found persistence of film processing in the equipment with directly adapted film processing unit according to DIN and ONORM. The checking of film persistence as demanded by DIN 6868 in these equipment could therefore be performed in longer periods. Systems with conventional darkroom processing comparatively show plain increased fluctuation, and hence the demanded daily control is essential to guarantee appropriate reaction and constant quality of documentation.

  8. Deep supervised dictionary learning for no-reference image quality assessment

    NASA Astrophysics Data System (ADS)

    Huang, Yuge; Liu, Xuesong; Tian, Xiang; Zhou, Fan; Chen, Yaowu; Jiang, Rongxin

    2018-03-01

    We propose a deep convolutional neural network (CNN) for general no-reference image quality assessment (NR-IQA), i.e., accurate prediction of image quality without a reference image. The proposed model consists of three components such as a local feature extractor that is a fully CNN, an encoding module with an inherent dictionary that aggregates local features to output a fixed-length global quality-aware image representation, and a regression module that maps the representation to an image quality score. Our model can be trained in an end-to-end manner, and all of the parameters, including the weights of the convolutional layers, the dictionary, and the regression weights, are simultaneously learned from the loss function. In addition, the model can predict quality scores for input images of arbitrary sizes in a single step. We tested our method on commonly used image quality databases and showed that its performance is comparable with that of state-of-the-art general-purpose NR-IQA algorithms.

  9. Clinical image quality evaluation for panoramic radiography in Korean dental clinics

    PubMed Central

    Choi, Bo-Ram; Choi, Da-Hye; Huh, Kyung-Hoe; Yi, Won-Jin; Heo, Min-Suk; Choi, Soon-Chul; Bae, Kwang-Hak

    2012-01-01

    Purpose The purpose of this study was to investigate the level of clinical image quality of panoramic radiographs and to analyze the parameters that influence the overall image quality. Materials and Methods Korean dental clinics were asked to provide three randomly selected panoramic radiographs. An oral and maxillofacial radiology specialist evaluated those images using our self-developed Clinical Image Quality Evaluation Chart. Three evaluators classified the overall image quality of the panoramic radiographs and evaluated the causes of imaging errors. Results A total of 297 panoramic radiographs were collected from 99 dental hospitals and clinics. The mean of the scores according to the Clinical Image Quality Evaluation Chart was 79.9. In the classification of the overall image quality, 17 images were deemed 'optimal for obtaining diagnostic information,' 153 were 'adequate for diagnosis,' 109 were 'poor but diagnosable,' and nine were 'unrecognizable and too poor for diagnosis'. The results of the analysis of the causes of the errors in all the images are as follows: 139 errors in the positioning, 135 in the processing, 50 from the radiographic unit, and 13 due to anatomic abnormality. Conclusion Panoramic radiographs taken at local dental clinics generally have a normal or higher-level image quality. Principal factors affecting image quality were positioning of the patient and image density, sharpness, and contrast. Therefore, when images are taken, the patient position should be adjusted with great care. Also, standardizing objective criteria of image density, sharpness, and contrast is required to evaluate image quality effectively. PMID:23071969

  10. Reliability and validity of the body image quality of life inventory: version for Brazilian burn victims.

    PubMed

    Assunção, Flávia Fernanda Oliveira; Dantas, Rosana Aparecida Spadoti; Ciol, Márcia Aparecida; Gonçalves, Natália; Farina, Jayme Adriano; Rossi, Lidia Aparecida

    2013-06-01

    The aims of this study were to adapt the Body Image Quality of Life Inventory (BIQLI) into Brazilian Portuguese (BP) and to assess the psychometric properties of the adapted version. Construct validity was assessed by correlating the BIQLI-BP scores with the Rosenberg's Self-Esteem Scale, with Burns Specific Health Scale-Revised (BSHS-R), and with gender, total body surface area burned, and visibility of the scars. Participants were 77 adult burn patients. Cronbach's alpha for the adapted version was .90 and moderate linear correlations were found between body image and self-esteem and between BIQLI-BP scores and two domains of the BSHS-R: affect and body image and interpersonal relationships. The BIQLI-BP showed acceptable levels of reliability and validity for Brazilian burn patients. Copyright © 2013 Wiley Periodicals, Inc.

  11. Image quality prediction - An aid to the Viking lander imaging investigation on Mars

    NASA Technical Reports Server (NTRS)

    Huck, F. O.; Wall, S. D.

    1976-01-01

    Image quality criteria and image quality predictions are formulated for the multispectral panoramic cameras carried by the Viking Mars landers. Image quality predictions are based on expected camera performance, Mars surface radiance, and lighting and viewing geometry (fields of view, Mars lander shadows, solar day-night alternation), and are needed in diagnosis of camera performance, in arriving at a preflight imaging strategy, and revision of that strategy should the need arise. Landing considerations, camera control instructions, camera control logic, aspects of the imaging process (spectral response, spatial response, sensitivity), and likely problems are discussed. Major concerns include: degradation of camera response by isotope radiation, uncertainties in lighting and viewing geometry and in landing site local topography, contamination of camera window by dust abrasion, and initial errors in assigning camera dynamic ranges (gains and offsets).

  12. Image quality assessment for CT used on small animals

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Cisneros, Isabela Paredes, E-mail: iparedesc@unal.edu.co; Agulles-Pedrós, Luis, E-mail: lagullesp@unal.edu.co

    Image acquisition on a CT scanner is nowadays necessary in almost any kind of medical study. Its purpose, to produce anatomical images with the best achievable quality, implies the highest diagnostic radiation exposure to patients. Image quality can be measured quantitatively based on parameters such as noise, uniformity and resolution. This measure allows the determination of optimal parameters of operation for the scanner in order to get the best diagnostic image. A human Phillips CT scanner is the first one minded for veterinary-use exclusively in Colombia. The aim of this study was to measure the CT image quality parameters usingmore » an acrylic phantom and then, using the computational tool MATLAB, determine these parameters as a function of current value and window of visualization, in order to reduce dose delivery by keeping the appropriate image quality.« less

  13. Image quality assessment for CT used on small animals

    NASA Astrophysics Data System (ADS)

    Cisneros, Isabela Paredes; Agulles-Pedrós, Luis

    2016-07-01

    Image acquisition on a CT scanner is nowadays necessary in almost any kind of medical study. Its purpose, to produce anatomical images with the best achievable quality, implies the highest diagnostic radiation exposure to patients. Image quality can be measured quantitatively based on parameters such as noise, uniformity and resolution. This measure allows the determination of optimal parameters of operation for the scanner in order to get the best diagnostic image. A human Phillips CT scanner is the first one minded for veterinary-use exclusively in Colombia. The aim of this study was to measure the CT image quality parameters using an acrylic phantom and then, using the computational tool MatLab, determine these parameters as a function of current value and window of visualization, in order to reduce dose delivery by keeping the appropriate image quality.

  14. Evaluating imaging quality between different ghost imaging systems based on the coherent-mode representation

    NASA Astrophysics Data System (ADS)

    Shen, Qian; Bai, Yanfeng; Shi, Xiaohui; Nan, Suqin; Qu, Lijie; Li, Hengxing; Fu, Xiquan

    2017-07-01

    The difference in imaging quality between different ghost imaging schemes is studied by using coherent-mode representation of partially coherent fields. It is shown that the difference mainly relies on the distribution changes of the decomposition coefficients of the object imaged when the light source is fixed. For a new-designed imaging scheme, we only need to give the distribution of the decomposition coefficients and compare them with that of the existing imaging system, thus one can predict imaging quality. By choosing several typical ghost imaging systems, we theoretically and experimentally verify our results.

  15. Subjective evaluation of compressed image quality

    NASA Astrophysics Data System (ADS)

    Lee, Heesub; Rowberg, Alan H.; Frank, Mark S.; Choi, Hyung-Sik; Kim, Yongmin

    1992-05-01

    Lossy data compression generates distortion or error on the reconstructed image and the distortion becomes visible as the compression ratio increases. Even at the same compression ratio, the distortion appears differently depending on the compression method used. Because of the nonlinearity of the human visual system and lossy data compression methods, we have evaluated subjectively the quality of medical images compressed with two different methods, an intraframe and interframe coding algorithms. The evaluated raw data were analyzed statistically to measure interrater reliability and reliability of an individual reader. Also, the analysis of variance was used to identify which compression method is better statistically, and from what compression ratio the quality of a compressed image is evaluated as poorer than that of the original. Nine x-ray CT head images from three patients were used as test cases. Six radiologists participated in reading the 99 images (some were duplicates) compressed at four different compression ratios, original, 5:1, 10:1, and 15:1. The six readers agree more than by chance alone and their agreement was statistically significant, but there were large variations among readers as well as within a reader. The displacement estimated interframe coding algorithm is significantly better in quality than that of the 2-D block DCT at significance level 0.05. Also, 10:1 compressed images with the interframe coding algorithm do not show any significant differences from the original at level 0.05.

  16. Impact of image quality on OCT angiography based quantitative measurements.

    PubMed

    Al-Sheikh, Mayss; Ghasemi Falavarjani, Khalil; Akil, Handan; Sadda, SriniVas R

    2017-01-01

    To study the impact of image quality on quantitative measurements and the frequency of segmentation error with optical coherence tomography angiography (OCTA). Seventeen eyes of 10 healthy individuals were included in this study. OCTA was performed using a swept-source device (Triton, Topcon). Each subject underwent three scanning sessions 1-2 min apart; the first two scans were obtained under standard conditions and for the third session, the image quality index was reduced using application of a topical ointment. En face OCTA images of the retinal vasculature were generated using the default segmentation for the superficial and deep retinal layer (SRL, DRL). Intraclass correlation coefficient (ICC) was used as a measure for repeatability. The frequency of segmentation error, motion artifact, banding artifact and projection artifact was also compared among the three sessions. The frequency of segmentation error, and motion artifact was statistically similar between high and low image quality sessions (P = 0.707, and P = 1 respectively). However, the frequency of projection and banding artifact was higher with a lower image quality. The vessel density in the SRL was highly repeatable in the high image quality sessions (ICC = 0.8), however, the repeatability was low, comparing the high and low image quality measurements (ICC = 0.3). In the DRL, the repeatability of the vessel density measurements was fair in the high quality sessions (ICC = 0.6 and ICC = 0.5, with and without automatic artifact removal, respectively) and poor comparing high and low image quality sessions (ICC = 0.3 and ICC = 0.06, with and without automatic artifact removal, respectively). The frequency of artifacts is higher and the repeatability of the measurements is lower with lower image quality. The impact of image quality index should be always considered in OCTA based quantitative measurements.

  17. Review of Image Quality Measures for Solar Imaging

    NASA Astrophysics Data System (ADS)

    Popowicz, Adam; Radlak, Krystian; Bernacki, Krzysztof; Orlov, Valeri

    2017-12-01

    Observations of the solar photosphere from the ground encounter significant problems caused by Earth's turbulent atmosphere. Before image reconstruction techniques can be applied, the frames obtained in the most favorable atmospheric conditions (the so-called lucky frames) have to be carefully selected. However, estimating the quality of images containing complex photospheric structures is not a trivial task, and the standard routines applied in nighttime lucky imaging observations are not applicable. In this paper we evaluate 36 methods dedicated to the assessment of image quality, which were presented in the literature over the past 40 years. We compare their effectiveness on simulated solar observations of both active regions and granulation patches, using reference data obtained by the Solar Optical Telescope on the Hinode satellite. To create images that are affected by a known degree of atmospheric degradation, we employed the random wave vector method, which faithfully models all the seeing characteristics. The results provide useful information about the method performances, depending on the average seeing conditions expressed by the ratio of the telescope's aperture to the Fried parameter, D/r0. The comparison identifies three methods for consideration by observers: Helmli and Scherer's mean, the median filter gradient similarity, and the discrete cosine transform energy ratio. While the first method requires less computational effort and can be used effectively in virtually any atmospheric conditions, the second method shows its superiority at good seeing (D/r0<4). The third method should mainly be considered for the post-processing of strongly blurred images.

  18. Dynamic flat panel detector versus image intensifier in cardiac imaging: dose and image quality

    NASA Astrophysics Data System (ADS)

    Vano, E.; Geiger, B.; Schreiner, A.; Back, C.; Beissel, J.

    2005-12-01

    The practical aspects of the dosimetric and imaging performance of a digital x-ray system for cardiology procedures were evaluated. The system was configured with an image intensifier (II) and later upgraded to a dynamic flat panel detector (FD). Entrance surface air kerma (ESAK) to phantoms of 16, 20, 24 and 28 cm of polymethyl methacrylate (PMMA) and the image quality of a test object were measured. Images were evaluated directly on the monitor and with numerical methods (noise and signal-to-noise ratio). Information contained in the DICOM header for dosimetry audit purposes was also tested. ESAK values per frame (or kerma rate) for the most commonly used cine and fluoroscopy modes for different PMMA thicknesses and for field sizes of 17 and 23 cm for II, and 20 and 25 cm for FD, produced similar results in the evaluated system with both technologies, ranging between 19 and 589 µGy/frame (cine) and 5 and 95 mGy min-1 (fluoroscopy). Image quality for these dose settings was better for the FD version. The 'study dosimetric report' is comprehensive, and its numerical content is sufficiently accurate. There is potential in the future to set those systems with dynamic FD to lower doses than are possible in the current II versions, especially for digital cine runs, or to benefit from improved image quality.

  19. Readout-Segmented Echo-Planar Imaging in Diffusion-Weighted MR Imaging in Breast Cancer: Comparison with Single-Shot Echo-Planar Imaging in Image Quality

    PubMed Central

    Kim, Yun Ju; Kang, Bong Joo; Park, Chang Suk; Kim, Hyeon Sook; Son, Yo Han; Porter, David Andrew; Song, Byung Joo

    2014-01-01

    Objective The purpose of this study was to compare the image quality of standard single-shot echo-planar imaging (ss-EPI) and that of readout-segmented EPI (rs-EPI) in patients with breast cancer. Materials and Methods Seventy-one patients with 74 breast cancers underwent both ss-EPI and rs-EPI. For qualitative comparison of image quality, three readers independently assessed the two sets of diffusion-weighted (DW) images. To evaluate geometric distortion, a comparison was made between lesion lengths derived from contrast enhanced MR (CE-MR) images and those obtained from the corresponding DW images. For assessment of image parameters, signal-to-noise ratio (SNR), lesion contrast, and contrast-to-noise ratio (CNR) were calculated. Results The rs-EPI was superior to ss-EPI in most criteria regarding the qualitative image quality. Anatomical structure distinction, delineation of the lesion, ghosting artifact, and overall image quality were significantly better in rs-EPI. Regarding the geometric distortion, lesion length on ss-EPI was significantly different from that of CE-MR, whereas there were no significant differences between CE-MR and rs-EPI. The rs-EPI was superior to ss-EPI in SNR and CNR. Conclusion Readout-segmented EPI is superior to ss-EPI in the aspect of image quality in DW MR imaging of the breast. PMID:25053898

  20. Topological image texture analysis for quality assessment

    NASA Astrophysics Data System (ADS)

    Asaad, Aras T.; Rashid, Rasber Dh.; Jassim, Sabah A.

    2017-05-01

    Image quality is a major factor influencing pattern recognition accuracy and help detect image tampering for forensics. We are concerned with investigating topological image texture analysis techniques to assess different type of degradation. We use Local Binary Pattern (LBP) as a texture feature descriptor. For any image construct simplicial complexes for selected groups of uniform LBP bins and calculate persistent homology invariants (e.g. number of connected components). We investigated image quality discriminating characteristics of these simplicial complexes by computing these models for a large dataset of face images that are affected by the presence of shadows as a result of variation in illumination conditions. Our tests demonstrate that for specific uniform LBP patterns, the number of connected component not only distinguish between different levels of shadow effects but also help detect the infected regions as well.

  1. Image Quality Performance Measurement of the microPET Focus 120

    NASA Astrophysics Data System (ADS)

    Ballado, Fernando Trejo; López, Nayelli Ortega; Flores, Rafael Ojeda; Ávila-Rodríguez, Miguel A.

    2010-12-01

    The aim of this work is to evaluate the characteristics involved in the image reconstruction of the microPET Focus 120. For this evaluation were used two different phantoms; a miniature hot-rod Derenzo phantom and a National Electrical Manufacturers Association (NEMA) NU4-2008 image quality (IQ) phantom. The best image quality was obtained when using OSEM3D as the reconstruction method reaching a spatial resolution of 1.5 mm with the Derenzo phantom filled with 18F. Image quality test results indicate a superior image quality for the Focus 120 when compared to previous microPET models.

  2. Effect of sorghum flour addition on in vitro starch digestibility, cooking quality, and consumer acceptability of durum wheat pasta.

    PubMed

    Khan, Imran; Yousif, Adel M; Johnson, Stuart K; Gamlath, Shirani

    2014-08-01

    Whole grain sorghum is a valuable source of resistant starch and polyphenolic antioxidants and its addition into staple food like pasta may reduce the starch digestibility. However, incorporating nondurum wheat materials into pasta provides a challenge in terms of maintaining cooking quality and consumer acceptability. Pasta was prepared from 100% durum wheat semolina (DWS) as control or by replacing DWS with either wholegrain red sorghum flour (RSF) or white sorghum flour (WSF) each at 20%, 30%, and 40% incorporation levels, following a laboratory-scale procedure. Pasta samples were evaluated for proximate composition, in vitro starch digestibility, cooking quality, and consumer acceptability. The addition of both RSF and WSF lowered the extent of in vitro starch digestion at all substitution levels compared to the control pasta. The rapidly digestible starch was lowered in all the sorghum-containing pastas compared to the control pasta. Neither RSF or WSF addition affected the pasta quality attributes (water absorption, swelling index, dry matter, adhesiveness, cohesiveness, and springiness), except color and hardness which were negatively affected. Consumer sensory results indicated that pasta samples containing 20% and 30% RSF or WSF had acceptable palatability based on meeting one or both of the preset acceptability criteria. It is concluded that the addition of wholegrain sorghum flour to pasta at 30% incorporation level is possible to reduce starch digestibility, while maintaining adequate cooking quality and consumer acceptability. © 2014 Institute of Food Technologists®

  3. Toward a perceptual image quality assessment of color quantized images

    NASA Astrophysics Data System (ADS)

    Frackiewicz, Mariusz; Palus, Henryk

    2018-04-01

    Color image quantization is an important operation in the field of color image processing. In this paper, we consider new perceptual image quality metrics for assessment of quantized images. These types of metrics, e.g. DSCSI, MDSIs, MDSIm and HPSI achieve the highest correlation coefficients with MOS during tests on the six publicly available image databases. Research was limited to images distorted by two types of compression: JPG and JPG2K. Statistical analysis of correlation coefficients based on the Friedman test and post-hoc procedures showed that the differences between the four new perceptual metrics are not statistically significant.

  4. A study of image quality for radar image processing. [synthetic aperture radar imagery

    NASA Technical Reports Server (NTRS)

    King, R. W.; Kaupp, V. H.; Waite, W. P.; Macdonald, H. C.

    1982-01-01

    Methods developed for image quality metrics are reviewed with focus on basic interpretation or recognition elements including: tone or color; shape; pattern; size; shadow; texture; site; association or context; and resolution. Seven metrics are believed to show promise as a way of characterizing the quality of an image: (1) the dynamic range of intensities in the displayed image; (2) the system signal-to-noise ratio; (3) the system spatial bandwidth or bandpass; (4) the system resolution or acutance; (5) the normalized-mean-square-error as a measure of geometric fidelity; (6) the perceptual mean square error; and (7) the radar threshold quality factor. Selective levels of degradation are being applied to simulated synthetic radar images to test the validity of these metrics.

  5. The use of the general image quality equation in the design and evaluation of imaging systems

    NASA Astrophysics Data System (ADS)

    Cota, Steve A.; Florio, Christopher J.; Duvall, David J.; Leon, Michael A.

    2009-08-01

    The design of any modern imaging system is the end result of many trade studies, each seeking to optimize image quality within real world constraints such as cost, schedule and overall risk. The National Imagery Interpretability Rating Scale (NIIRS) is a useful measure of image quality, because, by characterizing the overall interpretability of an image, it combines into one metric those contributors to image quality to which a human interpreter is most sensitive. The main drawback to using a NIIRS rating as a measure of image quality in engineering trade studies is the fact that it is tied to the human observer and cannot be predicted from physical principles and engineering parameters alone. The General Image Quality Equation (GIQE) of Leachtenauer et al. 1997 [Appl. Opt. 36, 8322-8328 (1997)] is a regression of actual image analyst NIIRS ratings vs. readily calculable engineering metrics, and provides a mechanism for using the expected NIIRS rating of an imaging system in the design and evaluation process. In this paper, we will discuss how we use the GIQE in conjunction with The Aerospace Corporation's Parameterized Image Chain Analysis & Simulation SOftware (PICASSO) to evaluate imager designs, taking a hypothetical high resolution commercial imaging system as an example.

  6. Recognizable or Not: Towards Image Semantic Quality Assessment for Compression

    NASA Astrophysics Data System (ADS)

    Liu, Dong; Wang, Dandan; Li, Houqiang

    2017-12-01

    Traditionally, image compression was optimized for the pixel-wise fidelity or the perceptual quality of the compressed images given a bit-rate budget. But recently, compressed images are more and more utilized for automatic semantic analysis tasks such as recognition and retrieval. For these tasks, we argue that the optimization target of compression is no longer perceptual quality, but the utility of the compressed images in the given automatic semantic analysis task. Accordingly, we propose to evaluate the quality of the compressed images neither at pixel level nor at perceptual level, but at semantic level. In this paper, we make preliminary efforts towards image semantic quality assessment (ISQA), focusing on the task of optical character recognition (OCR) from compressed images. We propose a full-reference ISQA measure by comparing the features extracted from text regions of original and compressed images. We then propose to integrate the ISQA measure into an image compression scheme. Experimental results show that our proposed ISQA measure is much better than PSNR and SSIM in evaluating the semantic quality of compressed images; accordingly, adopting our ISQA measure to optimize compression for OCR leads to significant bit-rate saving compared to using PSNR or SSIM. Moreover, we perform subjective test about text recognition from compressed images, and observe that our ISQA measure has high consistency with subjective recognizability. Our work explores new dimensions in image quality assessment, and demonstrates promising direction to achieve higher compression ratio for specific semantic analysis tasks.

  7. Quality of life and acceptability of medical versus surgical management of early pregnancy failure*

    PubMed Central

    Harwood, B; Nansel, T

    2008-01-01

    Objective This study compares quality of life (QOL) and acceptability of medical versus surgical treatment of early pregnancy failure (EPF). Design A randomised clinical trial of treatment for EPF compared misoprostol vaginally versus vacuum aspiration (VA). Setting A multisite trial at four US Urban University Hospitals. Population A total of 652 women with an EPF were randomised to treatment. Methods Participants completed a daily symptom diary and a questionnaire 2 weeks after treatment. Main outcome measures The questionnaire assessment included subscales of the Short Form-36 Health Survey Revised for QOL and measures of wellbeing, recovery difficulties, and treatment acceptability. Results The two groups did not differ in mean scores for QOL except bodily pain; medical treatment was associated with higher levels of bodily pain than VA (P < 0.001). Success of treatment was not related to QOL, but acceptability of the procedure was decreased for medical therapy if unsuccessful (P = 0.003). Type of treatment was not associated with differences in recovery, and the two groups reported similar acceptability except for cramping (P = 0.02), bleeding (P < 0.001), and symptom duration (P = 0.03). Conclusions Despite reporting greater pain and lower acceptability of treatment-related symptoms, QOL and treatment acceptability were similar for medical and surgical treatment of EPF. Acceptability, but not QOL, was influenced by success or failure of medical management. PMID:18271887

  8. Image-Processing Techniques for the Creation of Presentation-Quality Astronomical Images

    NASA Astrophysics Data System (ADS)

    Rector, Travis A.; Levay, Zoltan G.; Frattare, Lisa M.; English, Jayanne; Pu'uohau-Pummill, Kirk

    2007-02-01

    The quality of modern astronomical data and the agility of current image-processing software enable the visualization of data in a way that exceeds the traditional definition of an astronomical image. Two developments in particular have led to a fundamental change in how astronomical images can be assembled. First, the availability of high-quality multiwavelength and narrowband data allow for images that do not correspond to the wavelength sensitivity of the human eye, thereby introducing ambiguity in the usage and interpretation of color. Second, many image-processing software packages now use a layering metaphor that allows for any number of astronomical data sets to be combined into a color image. With this technique, images with as many as eight data sets have been produced. Each data set is intensity-scaled and colorized independently, creating an immense parameter space that can be used to assemble the image. Since such images are intended for data visualization, scaling and color schemes must be chosen that best illustrate the science. A practical guide is presented on how to use the layering metaphor to generate publication-ready astronomical images from as many data sets as desired. A methodology is also given on how to use intensity scaling, color, and composition to create contrasts in an image that highlight the scientific detail. Examples of image creation are discussed.

  9. Image quality comparison of two adaptive statistical iterative reconstruction (ASiR, ASiR-V) algorithms and filtered back projection in routine liver CT.

    PubMed

    Chen, Li-Hong; Jin, Chao; Li, Jian-Ying; Wang, Ge-Liang; Jia, Yong-Jun; Duan, Hai-Feng; Pan, Ning; Guo, Jianxin

    2018-06-06

    To compare image quality of two adaptive statistical iterative reconstruction (ASiR and ASiR-V) algorithms using objective and subjective metrics for routine liver CT, with the conventional filtered back projection (FBP) reconstructions as reference standards. This institutional review board-approved study included 52 patients with clinically suspected hepatic metastases. Patients were divided equally into ASiR and ASiR-V groups with same scan parameters. Images were reconstructed with ASiR and ASiR-V from 0 (FBP) to 100% blending percentages at 10% interval in its respective group. Mean and standard deviation of CT numbers for liver parenchyma were recorded. Two experienced radiologists reviewed all images for image quality blindly and independently. Data were statistically analyzed. There was no difference in CT dose index between ASiR and ASiR-V groups. As the percentage of ASiR and ASiR-V increased from 10 to 100% , image noise reduced by 8.6 -57.9% and 8.9-81.6%, respectively, compared with FBP. There was substantial interobserver agreement in image quality assessment for ASiR and ASiR-V images. Compared with FBP reconstruction, subjective image quality scores of ASiR and ASiR-V improved significantly as percentage increased from 10 to 80% for ASiR (peaked at 50% with 32.2% noise reduction) and from 10 to 90% (peaked at 60% with 51.5% noise reduction) for ASiR-V. Both ASiR and ASiR-V improved the objective and subjective image quality for routine liver CT compared with FBP. ASiR-V provided further image quality improvement with higher acceptable percentage than ASiR, and ASiR-V60% had the highest image quality score. Advances in knowledge: (1) Both ASiR and ASiR-V significantly reduce image noise compared with conventional FBP reconstruction. (2) ASiR-V with 60 blending percentage provides the highest image quality score in routine liver CT.

  10. Assessing microscope image focus quality with deep learning.

    PubMed

    Yang, Samuel J; Berndl, Marc; Michael Ando, D; Barch, Mariya; Narayanaswamy, Arunachalam; Christiansen, Eric; Hoyer, Stephan; Roat, Chris; Hung, Jane; Rueden, Curtis T; Shankar, Asim; Finkbeiner, Steven; Nelson, Philip

    2018-03-15

    Large image datasets acquired on automated microscopes typically have some fraction of low quality, out-of-focus images, despite the use of hardware autofocus systems. Identification of these images using automated image analysis with high accuracy is important for obtaining a clean, unbiased image dataset. Complicating this task is the fact that image focus quality is only well-defined in foreground regions of images, and as a result, most previous approaches only enable a computation of the relative difference in quality between two or more images, rather than an absolute measure of quality. We present a deep neural network model capable of predicting an absolute measure of image focus on a single image in isolation, without any user-specified parameters. The model operates at the image-patch level, and also outputs a measure of prediction certainty, enabling interpretable predictions. The model was trained on only 384 in-focus Hoechst (nuclei) stain images of U2OS cells, which were synthetically defocused to one of 11 absolute defocus levels during training. The trained model can generalize on previously unseen real Hoechst stain images, identifying the absolute image focus to within one defocus level (approximately 3 pixel blur diameter difference) with 95% accuracy. On a simpler binary in/out-of-focus classification task, the trained model outperforms previous approaches on both Hoechst and Phalloidin (actin) stain images (F-scores of 0.89 and 0.86, respectively over 0.84 and 0.83), despite only having been presented Hoechst stain images during training. Lastly, we observe qualitatively that the model generalizes to two additional stains, Hoechst and Tubulin, of an unseen cell type (Human MCF-7) acquired on a different instrument. Our deep neural network enables classification of out-of-focus microscope images with both higher accuracy and greater precision than previous approaches via interpretable patch-level focus and certainty predictions. The use of

  11. Roles of body image-related experiential avoidance and uncommitted living in the link between body image and women's quality of life.

    PubMed

    Trindade, Inês A; Ferreira, Cláudia; Pinto-Gouveia, José

    2018-01-01

    The current study aimed to test whether the associations of body mass index, body image discrepancy, and social comparison based on physical appearance with women's psychological quality of life (QoL) would be explained by the mechanisms of body image-related experiential avoidance and patterns of uncommitted living. The sample was collected from October 2014 to March 2015 and included 737 female college students (aged between 18 and 25 years) who completed validated self-report measures. Results demonstrated that the final path model explained 43% of psychological QoL and revealed an excellent fit. Body image-related experiential avoidance had a meditational role in the association between body image discrepancy and psychological QoL. Further, the link between social comparison based on physical appearance and psychological QoL was partially mediated by body image-related experiential avoidance and uncommitted living. These findings indicate that the key mechanisms of the relationship between body image and young women's QoL were those related to maladaptive emotion regulation. It thus seems that interventions aiming to promote mental health in this population should promote acceptance of internal experiences related to physical appearance (e.g., sensations, thoughts, or emotions) and the engagement in behaviors committed to life values.

  12. Assessment of CT image quality using a Bayesian approach

    NASA Astrophysics Data System (ADS)

    Reginatto, M.; Anton, M.; Elster, C.

    2017-08-01

    One of the most promising approaches for evaluating CT image quality is task-specific quality assessment. This involves a simplified version of a clinical task, e.g. deciding whether an image belongs to the class of images that contain the signature of a lesion or not. Task-specific quality assessment can be done by model observers, which are mathematical procedures that carry out the classification task. The most widely used figure of merit for CT image quality is the area under the ROC curve (AUC), a quantity which characterizes the performance of a given model observer. In order to estimate AUC from a finite sample of images, different approaches from classical statistics have been suggested. The goal of this paper is to introduce task-specific quality assessment of CT images to metrology and to propose a novel Bayesian estimation of AUC for the channelized Hotelling observer (CHO) applied to the task of detecting a lesion at a known image location. It is assumed that signal-present and signal-absent images follow multivariate normal distributions with the same covariance matrix. The Bayesian approach results in a posterior distribution for the AUC of the CHO which provides in addition a complete characterization of the uncertainty of this figure of merit. The approach is illustrated by its application to both simulated and experimental data.

  13. Predictors of acceptance of offered care management intervention services in a quality improvement trial for dementia.

    PubMed

    Kaisey, Marwa; Mittman, Brian; Pearson, Marjorie; Connor, Karen I; Chodosh, Joshua; Vassar, Stefanie D; Nguyen, France T; Vickrey, Barbara G

    2012-10-01

    Care management approaches have been proven to improve outcomes for patients with dementia and their family caregivers (dyads). However, acceptance of services in these programs is incomplete, impacting effectiveness. Acceptance may be related to dyad as well as healthcare system characteristics, but knowledge about factors associated with program acceptance is lacking. This study investigates patient, caregiver, and healthcare system characteristics associated with acceptance of offered care management services. This study analyzed data from the intervention arm of a cluster randomized controlled trial of a comprehensive dementia care management intervention. There were 408 patient-caregiver dyads enrolled in the study, of which 238 dyads were randomized to the intervention. Caregiver, patient, and health system factors associated with participation in offered care management services were assessed through bivariate and multivariate regression analyses. Out of the 238 dyads, 9 were ineligible for this analysis, leaving data of 229 dyads in this sample. Of these, 185 dyads accepted offered care management services, and 44 dyads did not. Multivariate analyses showed that higher likelihood of acceptance of care management services was uniquely associated with cohabitation of caregiver and patient (p < 0.001), lesser severity of dementia (p = 0.03), and higher patient comorbidity (p = 0.03); it also varied across healthcare organization sites. Understanding factors that influence care management participation could result in increased adoption of successful programs to improve quality of care. Using these factors to revise both program design as well as program promotion may also benefit external validity of future quality improvement research trials. Copyright © 2011 John Wiley & Sons, Ltd.

  14. Medical image enhancement using resolution synthesis

    NASA Astrophysics Data System (ADS)

    Wong, Tak-Shing; Bouman, Charles A.; Thibault, Jean-Baptiste; Sauer, Ken D.

    2011-03-01

    We introduce a post-processing approach to improve the quality of CT reconstructed images. The scheme is adapted from the resolution-synthesis (RS)1 interpolation algorithm. In this approach, we consider the input image, scanned at a particular dose level, as a degraded version of a high quality image scanned at a high dose level. Image enhancement is achieved by predicting the high quality image by classification based linear regression. To improve the robustness of our scheme, we also apply the minimum description length principle to determine the optimal number of predictors to use in the scheme, and the ridge regression to regularize the design of the predictors. Experimental results show that our scheme is effective in reducing the noise in images reconstructed from filtered back projection without significant loss of image details. Alternatively, our scheme can also be applied to reduce dose while maintaining image quality at an acceptable level.

  15. Perception of drinking water safety and factors influencing acceptance and sustainability of a water quality intervention in rural southern India.

    PubMed

    Francis, Mark Rohit; Nagarajan, Guru; Sarkar, Rajiv; Mohan, Venkata Raghava; Kang, Gagandeep; Balraj, Vinohar

    2015-07-30

    Acceptance and long-term sustainability of water quality interventions are pivotal to realizing continued health benefits. However, there is limited research attempting to understand the factors that influence compliance to or adoption of such interventions. Eight focus group discussions with parents of young children--including compliant and not compliant households participating in an intervention study, and three key-informant interviews with village headmen were conducted between April and May 2014 to understand perceptions on the effects of unsafe water on health, household drinking water treatment practices, and the factors influencing acceptance and sustainability of an ongoing water quality intervention in a rural population of southern India. The ability to recognize health benefits from the intervention, ease of access to water distribution centers and the willingness to pay for intervention maintenance were factors facilitating acceptance and sustainability of the water quality intervention. On the other hand, faulty perceptions on water treatment, lack of knowledge about health hazards associated with drinking unsafe water, false sense of protection from locally available water, resistance to change in taste or odor of water and a lack of support from male members of the household were important factors impeding acceptance and long term use of the intervention. This study highlights the need to effectively involve communities at important stages of implementation for long term success of water quality interventions. Timely research on the factors influencing uptake of water quality interventions prior to implementation will ensure greater acceptance and sustainability of such interventions in low income settings.

  16. High-pitch Helical Dual-source Computed Tomographic Pulmonary Angiography: Comparing Image Quality in Inspiratory Breath-hold and During Free Breathing.

    PubMed

    Ajlan, Amr M; Binzaqr, Salma; Jadkarim, Dalia A; Jamjoom, Lamia G; Leipsic, Jonathon

    2016-01-01

    The purpose of this study was to compare qualitative and quantitative image parameters of dual-source high-pitch helical computed tomographic pulmonary angiography (CTPA) in breath-holding (BH) versus free-breathing (FB) patients. Ninety-nine consented patients (61 female individuals; mean age±SD, 49±18.7 y) were randomized into BH (n=45) versus FB (n=54) high-pitch helical CTPA. Patient characteristics and CTPA radiation doses were analyzed. Two readers assessed for pulmonary embolism (PE), transient interruption of contrast, and respiratory and cardiac motion. The readers used a subjective 3-point scale to rate the pulmonary artery opacification and lung parenchymal appearance. A single reader assessed mean pulmonary artery signal intensity, noise, contrast, signal to noise ratio, and contrast to noise ratio. PE was diagnosed in 16% BH and 19% FB patients. CTPAs of both groups were of excellent or acceptable quality for PE evaluation and of similar mean radiation doses (1.3 mSv). Transient interruption of contrast was seen in 5/45 (11%) BH and 5/54 (9%) FB patients (not statistically significant, P=0.54). No statistically significant difference was noted in cardiac, diaphragmatic, and lung parenchymal motion. Lung parenchymal assessment was excellent in all cases, except for 5/54 (9%) motion-affected FB cases with acceptable quality (statistically significant, P=0.03). No CTPA was considered nondiagnostic by any of the readers. No objective image quality differences were noted between both groups (P>0.05). High-pitch helical CTPA acquired during BH or in FB yields comparable image quality for the diagnosis of PE and lung pathology, with low radiation exposure. Only a modest increase in lung parenchymal artifacts is encountered in FB high-pitch helical CTPA.

  17. Physicochemical characterization of pure persimmon juice: nutritional quality and food acceptability.

    PubMed

    González, Eva; Vegara, Salud; Martí, Nuria; Valero, Manuel; Saura, Domingo

    2015-03-01

    Technological process for production of non-astringent persimmon (Diospyros kaki Thunb. cv. "Rojo Brillante") juice was described. The degree of fruit ripening expressed as color index (CI) varied between 12.37 and 16.33. Persimmon juice was characterized by determining physicochemical quality parameters as yield, total soluble solids (TSS), pH, titratable acidity (TA), organic acids, and main sugars. A thermal treatment of 90 ºC for 10 s was effective in controlling naturally occurring microorganisms for at least 105 d of storage without significantly affecting production of soluble brown pigments (BPs) and 5-hydroxymethyl furfural (5-HMF), total phenolic compounds (TPC), antioxidant capacity and acceptability of juice by panelists. Storage time affected all and each of the above parameters, reducing BPs, TPC and antioxidant capacity but increasing 5-HMF content. Refrigerated storage enhanced the acceptability of the juices. This information may be used by the juice industry as a starting point for production of pure persimmon juices. © 2015 Institute of Food Technologists®

  18. Imaging quality analysis of multi-channel scanning radiometer

    NASA Astrophysics Data System (ADS)

    Fan, Hong; Xu, Wujun; Wang, Chengliang

    2008-03-01

    Multi-channel scanning radiometer, on boarding FY-2 geostationary meteorological satellite, plays a key role in remote sensing because of its wide field of view and continuous multi-spectral images acquirements. It is significant to evaluate image quality after performance parameters of the imaging system are validated. Several methods of evaluating imaging quality are discussed. Of these methods, the most fundamental is the MTF. The MTF of photoelectric scanning remote instrument, in the scanning direction, is the multiplication of optics transfer function (OTF), detector transfer function (DTF) and electronics transfer function (ETF). For image motion compensation, moving speed of scanning mirror should be considered. The optical MTF measurement is performed in both the EAST/WEST and NORTH/SOUTH direction, whose values are used for alignment purposes and are used to determine the general health of the instrument during integration and testing. Imaging systems cannot perfectly reproduce what they see and end up "blurring" the image. Many parts of the imaging system can cause blurring. Among these are the optical elements, the sampling of the detector itself, post-processing, or the earth's atmosphere for systems that image through it. Through theory calculation and actual measurement, it is proved that DTF and ETF are the main factors of system MTF and the imaging quality can satisfy the requirement of instrument design.

  19. Comprehensive model for predicting perceptual image quality of smart mobile devices.

    PubMed

    Gong, Rui; Xu, Haisong; Luo, M R; Li, Haifeng

    2015-01-01

    An image quality model for smart mobile devices was proposed based on visual assessments of several image quality attributes. A series of psychophysical experiments were carried out on two kinds of smart mobile devices, i.e., smart phones and tablet computers, in which naturalness, colorfulness, brightness, contrast, sharpness, clearness, and overall image quality were visually evaluated under three lighting environments via categorical judgment method for various application types of test images. On the basis of Pearson correlation coefficients and factor analysis, the overall image quality could first be predicted by its two constituent attributes with multiple linear regression functions for different types of images, respectively, and then the mathematical expressions were built to link the constituent image quality attributes with the physical parameters of smart mobile devices and image appearance factors. The procedure and algorithms were applicable to various smart mobile devices, different lighting conditions, and multiple types of images, and performance was verified by the visual data.

  20. Model-Based Referenceless Quality Metric of 3D Synthesized Images Using Local Image Description.

    PubMed

    Gu, Ke; Jakhetiya, Vinit; Qiao, Jun-Fei; Li, Xiaoli; Lin, Weisi; Thalmann, Daniel

    2017-07-28

    New challenges have been brought out along with the emerging of 3D-related technologies such as virtual reality (VR), augmented reality (AR), and mixed reality (MR). Free viewpoint video (FVV), due to its applications in remote surveillance, remote education, etc, based on the flexible selection of direction and viewpoint, has been perceived as the development direction of next-generation video technologies and has drawn a wide range of researchers' attention. Since FVV images are synthesized via a depth image-based rendering (DIBR) procedure in the "blind" environment (without reference images), a reliable real-time blind quality evaluation and monitoring system is urgently required. But existing assessment metrics do not render human judgments faithfully mainly because geometric distortions are generated by DIBR. To this end, this paper proposes a novel referenceless quality metric of DIBR-synthesized images using the autoregression (AR)-based local image description. It was found that, after the AR prediction, the reconstructed error between a DIBR-synthesized image and its AR-predicted image can accurately capture the geometry distortion. The visual saliency is then leveraged to modify the proposed blind quality metric to a sizable margin. Experiments validate the superiority of our no-reference quality method as compared with prevailing full-, reduced- and no-reference models.

  1. Applying image quality in cell phone cameras: lens distortion

    NASA Astrophysics Data System (ADS)

    Baxter, Donald; Goma, Sergio R.; Aleksic, Milivoje

    2009-01-01

    This paper describes the framework used in one of the pilot studies run under the I3A CPIQ initiative to quantify overall image quality in cell-phone cameras. The framework is based on a multivariate formalism which tries to predict overall image quality from individual image quality attributes and was validated in a CPIQ pilot program. The pilot study focuses on image quality distortions introduced in the optical path of a cell-phone camera, which may or may not be corrected in the image processing path. The assumption is that the captured image used is JPEG compressed and the cellphone camera is set to 'auto' mode. As the used framework requires that the individual attributes to be relatively perceptually orthogonal, in the pilot study, the attributes used are lens geometric distortion (LGD) and lateral chromatic aberrations (LCA). The goal of this paper is to present the framework of this pilot project starting with the definition of the individual attributes, up to their quantification in JNDs of quality, a requirement of the multivariate formalism, therefore both objective and subjective evaluations were used. A major distinction in the objective part from the 'DSC imaging world' is that the LCA/LGD distortions found in cell-phone cameras, rarely exhibit radial behavior, therefore a radial mapping/modeling cannot be used in this case.

  2. Objective Assessment of Image Quality VI: Imaging in Radiation Therapy

    PubMed Central

    Barrett, Harrison H.; Kupinski, Matthew A.; Müeller, Stefan; Halpern, Howard J.; Morris, John C.; Dwyer, Roisin

    2015-01-01

    Earlier work on Objective Assessment of Image Quality (OAIQ) focused largely on estimation or classification tasks in which the desired outcome of imaging is accurate diagnosis. This paper develops a general framework for assessing imaging quality on the basis of therapeutic outcomes rather than diagnostic performance. By analogy to Receiver Operating Characteristic (ROC) curves and their variants as used in diagnostic OAIQ, the method proposed here utilizes the Therapy Operating Characteristic or TOC curves, which are plots of the probability of tumor control vs. the probability of normal-tissue complications as the overall dose level of a radiotherapy treatment is varied. The proposed figure of merit is the area under the TOC curve, denoted AUTOC. This paper reviews an earlier exposition of the theory of TOC and AUTOC, which was specific to the assessment of image-segmentation algorithms, and extends it to other applications of imaging in external-beam radiation treatment as well as in treatment with internal radioactive sources. For each application, a methodology for computing the TOC is presented. A key difference between ROC and TOC is that the latter can be defined for a single patient rather than a population of patients. PMID:24200954

  3. Quality evaluation of pansharpened hyperspectral images generated using multispectral images

    NASA Astrophysics Data System (ADS)

    Matsuoka, Masayuki; Yoshioka, Hiroki

    2012-11-01

    Hyperspectral remote sensing can provide a smooth spectral curve of a target by using a set of higher spectral resolution detectors. The spatial resolution of the hyperspectral images, however, is generally much lower than that of multispectral images due to the lower energy of incident radiation. Pansharpening is an image-fusion technique that generates higher spatial resolution multispectral images by combining lower resolution multispectral images with higher resolution panchromatic images. In this study, higher resolution hyperspectral images were generated by pansharpening of simulated lower hyperspectral and higher multispectral data. Spectral and spatial qualities of pansharpened images, then, were accessed in relation to the spectral bands of multispectral images. Airborne hyperspectral data of AVIRIS was used in this study, and it was pansharpened using six methods. Quantitative evaluations of pansharpened image are achieved using two frequently used indices, ERGAS, and the Q index.

  4. Landsat image data quality studies

    NASA Technical Reports Server (NTRS)

    Schueler, C. F.; Salomonson, V. V.

    1985-01-01

    Preliminary results of the Landsat-4 Image Data Quality Analysis (LIDQA) program to characterize the data obtained using the Thematic Mapper (TM) instrument on board the Landsat-4 and Landsat-5 satellites are reported. TM design specifications were compared to the obtained data with respect to four criteria, including spatial resolution; geometric fidelity; information content; and image relativity to Multispectral Scanner (MSS) data. The overall performance of the TM was rated excellent despite minor instabilities and radiometric anomalies in the data. Spatial performance of the TM exceeded design specifications in terms of both image sharpness and geometric accuracy, and the image utility of the TM data was at least twice as high as MSS data. The separability of alfalfa and sugar beet fields in a TM image is demonstrated.

  5. 40 CFR 86.610-98 - Compliance with acceptable quality level and passing and failing criteria for Selective...

    Code of Federal Regulations, 2012 CFR

    2012-07-01

    ... level and passing and failing criteria for Selective Enforcement Audits. 86.610-98 Section 86.610-98... quality level and passing and failing criteria for Selective Enforcement Audits. (a) The prescribed acceptable quality level is 40 percent. (b) A failed vehicle is one whose final deteriorated test results...

  6. 40 CFR 86.610-98 - Compliance with acceptable quality level and passing and failing criteria for Selective...

    Code of Federal Regulations, 2010 CFR

    2010-07-01

    ... level and passing and failing criteria for Selective Enforcement Audits. 86.610-98 Section 86.610-98... quality level and passing and failing criteria for Selective Enforcement Audits. (a) The prescribed acceptable quality level is 40 percent. (b) A failed vehicle is one whose final deteriorated test results...

  7. 40 CFR 86.610-98 - Compliance with acceptable quality level and passing and failing criteria for Selective...

    Code of Federal Regulations, 2011 CFR

    2011-07-01

    ... level and passing and failing criteria for Selective Enforcement Audits. 86.610-98 Section 86.610-98... quality level and passing and failing criteria for Selective Enforcement Audits. (a) The prescribed acceptable quality level is 40 percent. (b) A failed vehicle is one whose final deteriorated test results...

  8. 40 CFR 86.610-98 - Compliance with acceptable quality level and passing and failing criteria for Selective...

    Code of Federal Regulations, 2013 CFR

    2013-07-01

    ... level and passing and failing criteria for Selective Enforcement Audits. 86.610-98 Section 86.610-98... quality level and passing and failing criteria for Selective Enforcement Audits. (a) The prescribed acceptable quality level is 40 percent. (b) A failed vehicle is one whose final deteriorated test results...

  9. Human visual system consistent quality assessment for remote sensing image fusion

    NASA Astrophysics Data System (ADS)

    Liu, Jun; Huang, Junyi; Liu, Shuguang; Li, Huali; Zhou, Qiming; Liu, Junchen

    2015-07-01

    Quality assessment for image fusion is essential for remote sensing application. Generally used indices require a high spatial resolution multispectral (MS) image for reference, which is not always readily available. Meanwhile, the fusion quality assessments using these indices may not be consistent with the Human Visual System (HVS). As an attempt to overcome this requirement and inconsistency, this paper proposes an HVS-consistent image fusion quality assessment index at the highest resolution without a reference MS image using Gaussian Scale Space (GSS) technology that could simulate the HVS. The spatial details and spectral information of original and fused images are first separated in GSS, and the qualities are evaluated using the proposed spatial and spectral quality index respectively. The overall quality is determined without a reference MS image by a combination of the proposed two indices. Experimental results on various remote sensing images indicate that the proposed index is more consistent with HVS evaluation compared with other widely used indices that may or may not require reference images.

  10. Digital radiography: optimization of image quality and dose using multi-frequency software.

    PubMed

    Precht, H; Gerke, O; Rosendahl, K; Tingberg, A; Waaler, D

    2012-09-01

    New developments in processing of digital radiographs (DR), including multi-frequency processing (MFP), allow optimization of image quality and radiation dose. This is particularly promising in children as they are believed to be more sensitive to ionizing radiation than adults. To examine whether the use of MFP software reduces the radiation dose without compromising quality at DR of the femur in 5-year-old-equivalent anthropomorphic and technical phantoms. A total of 110 images of an anthropomorphic phantom were imaged on a DR system (Canon DR with CXDI-50 C detector and MLT[S] software) and analyzed by three pediatric radiologists using Visual Grading Analysis. In addition, 3,500 images taken of a technical contrast-detail phantom (CDRAD 2.0) provide an objective image-quality assessment. Optimal image-quality was maintained at a dose reduction of 61% with MLT(S) optimized images. Even for images of diagnostic quality, MLT(S) provided a dose reduction of 88% as compared to the reference image. Software impact on image quality was found significant for dose (mAs), dynamic range dark region and frequency band. By optimizing image processing parameters, a significant dose reduction is possible without significant loss of image quality.

  11. Enhancement of digital radiography image quality using a convolutional neural network.

    PubMed

    Sun, Yuewen; Li, Litao; Cong, Peng; Wang, Zhentao; Guo, Xiaojing

    2017-01-01

    Digital radiography system is widely used for noninvasive security check and medical imaging examination. However, the system has a limitation of lower image quality in spatial resolution and signal to noise ratio. In this study, we explored whether the image quality acquired by the digital radiography system can be improved with a modified convolutional neural network to generate high-resolution images with reduced noise from the original low-quality images. The experiment evaluated on a test dataset, which contains 5 X-ray images, showed that the proposed method outperformed the traditional methods (i.e., bicubic interpolation and 3D block-matching approach) as measured by peak signal to noise ratio (PSNR) about 1.3 dB while kept highly efficient processing time within one second. Experimental results demonstrated that a residual to residual (RTR) convolutional neural network remarkably improved the image quality of object structural details by increasing the image resolution and reducing image noise. Thus, this study indicated that applying this RTR convolutional neural network system was useful to improve image quality acquired by the digital radiography system.

  12. [Dry view laser imager--a new economical photothermal imaging method].

    PubMed

    Weberling, R

    1996-11-01

    The production of hard copies is currently achieved by means of laser imagers and wet film processing in systems attached either directly in or to the laser imager or in a darkroom. Variations in image quality resulting from a not always optimal wet film development are frequent. A newly developed thermographic film developer for laser films without liquid powdered chemicals, on the other hand, is environmentally preferable and reducing operating costs. The completely dry developing process provides permanent image documentation meeting the quality and safety requirements of RöV and BAK. One of the currently available systems of this type, the DryView Laser Imager is inexpensive and easy to install. The selective connection principle of the DryView Laser Imager can be expanded as required and accepts digital and/or analog interfaces with all imaging systems (CT, MR, DR, US, NM) from the various manufactures.

  13. High-quality JPEG compression history detection for fake uncompressed images

    NASA Astrophysics Data System (ADS)

    Zhang, Rong; Wang, Rang-Ding; Guo, Li-Jun; Jiang, Bao-Chuan

    2017-05-01

    Authenticity is one of the most important evaluation factors of images for photography competitions or journalism. Unusual compression history of an image often implies the illicit intent of its author. Our work aims at distinguishing real uncompressed images from fake uncompressed images that are saved in uncompressed formats but have been previously compressed. To detect the potential image JPEG compression, we analyze the JPEG compression artifacts based on the tetrolet covering, which corresponds to the local image geometrical structure. Since the compression can alter the structure information, the tetrolet covering indexes may be changed if a compression is performed on the test image. Such changes can provide valuable clues about the image compression history. To be specific, the test image is first compressed with different quality factors to generate a set of temporary images. Then, the test image is compared with each temporary image block-by-block to investigate whether the tetrolet covering index of each 4×4 block is different between them. The percentages of the changed tetrolet covering indexes corresponding to the quality factors (from low to high) are computed and used to form the p-curve, the local minimum of which may indicate the potential compression. Our experimental results demonstrate the advantage of our method to detect JPEG compressions of high quality, even the highest quality factors such as 98, 99, or 100 of the standard JPEG compression, from uncompressed-format images. At the same time, our detection algorithm can accurately identify the corresponding compression quality factor.

  14. Comparison of image quality and radiation dose between fixed tube current and combined automatic tube current modulation in craniocervical CT angiography.

    PubMed

    Lee, E J; Lee, S K; Agid, R; Howard, P; Bae, J M; terBrugge, K

    2009-10-01

    The combined automatic tube current modulation (ATCM) technique adapts and modulates the x-ray tube current in the x-y-z axis according to the patient's individual anatomy. We compared image quality and radiation dose of the combined ATCM technique with those of a fixed tube current (FTC) technique in craniocervical CT angiography performed with a 64-section multidetector row CT (MDCT) system. A retrospective review of craniocervical CT angiograms (CTAs) by using combined ATCM (n = 25) and FTC techniques (n = 25) was performed. Other CTA parameters, such as kilovolt (peak), matrix size, FOV, section thickness, pitch, contrast agent, and contrast injection techniques, were held constant. We recorded objective image noise in the muscles at 2 anatomic levels: radiation exposure doses (CT dose index volume and dose-length product); and subjective image quality parameters, such as vascular delineation of various arterial vessels, visibility of small arterial detail, image artifacts, and certainty of diagnosis. The Mann-Whitney U test was used for statistical analysis. No significant difference was detected in subjective image quality parameters between the FTC and combined ATCM techniques. Most subjects in both study groups (49/50, 98%) had acceptable subjective artifacts. The objective image noise values at shoulder level did not show a significant difference, but the noise value at the upper neck was higher with the combined ATCM (P < .05) technique. Significant reduction in radiation dose (18% reduction) was noted with the combined ATCM technique (P < .05). The combined ATCM technique for craniocervical CTA performed at 64-section MDCT substantially reduced radiation exposure dose but maintained diagnostic image quality.

  15. Gradient Magnitude Similarity Deviation: A Highly Efficient Perceptual Image Quality Index.

    PubMed

    Xue, Wufeng; Zhang, Lei; Mou, Xuanqin; Bovik, Alan C

    2014-02-01

    It is an important task to faithfully evaluate the perceptual quality of output images in many applications, such as image compression, image restoration, and multimedia streaming. A good image quality assessment (IQA) model should not only deliver high quality prediction accuracy, but also be computationally efficient. The efficiency of IQA metrics is becoming particularly important due to the increasing proliferation of high-volume visual data in high-speed networks. We present a new effective and efficient IQA model, called gradient magnitude similarity deviation (GMSD). The image gradients are sensitive to image distortions, while different local structures in a distorted image suffer different degrees of degradations. This motivates us to explore the use of global variation of gradient based local quality map for overall image quality prediction. We find that the pixel-wise gradient magnitude similarity (GMS) between the reference and distorted images combined with a novel pooling strategy-the standard deviation of the GMS map-can predict accurately perceptual image quality. The resulting GMSD algorithm is much faster than most state-of-the-art IQA methods, and delivers highly competitive prediction accuracy. MATLAB source code of GMSD can be downloaded at http://www4.comp.polyu.edu.hk/~cslzhang/IQA/GMSD/GMSD.htm.

  16. Blind CT image quality assessment via deep learning strategy: initial study

    NASA Astrophysics Data System (ADS)

    Li, Sui; He, Ji; Wang, Yongbo; Liao, Yuting; Zeng, Dong; Bian, Zhaoying; Ma, Jianhua

    2018-03-01

    Computed Tomography (CT) is one of the most important medical imaging modality. CT images can be used to assist in the detection and diagnosis of lesions and to facilitate follow-up treatment. However, CT images are vulnerable to noise. Actually, there are two major source intrinsically causing the CT data noise, i.e., the X-ray photo statistics and the electronic noise background. Therefore, it is necessary to doing image quality assessment (IQA) in CT imaging before diagnosis and treatment. Most of existing CT images IQA methods are based on human observer study. However, these methods are impractical in clinical for their complex and time-consuming. In this paper, we presented a blind CT image quality assessment via deep learning strategy. A database of 1500 CT images is constructed, containing 300 high-quality images and 1200 corresponding noisy images. Specifically, the high-quality images were used to simulate the corresponding noisy images at four different doses. Then, the images are scored by the experienced radiologists by the following attributes: image noise, artifacts, edge and structure, overall image quality, and tumor size and boundary estimation with five-point scale. We trained a network for learning the non-liner map from CT images to subjective evaluation scores. Then, we load the pre-trained model to yield predicted score from the test image. To demonstrate the performance of the deep learning network in IQA, correlation coefficients: Pearson Linear Correlation Coefficient (PLCC) and Spearman Rank Order Correlation Coefficient (SROCC) are utilized. And the experimental result demonstrate that the presented deep learning based IQA strategy can be used in the CT image quality assessment.

  17. A database for spectral image quality

    NASA Astrophysics Data System (ADS)

    Le Moan, Steven; George, Sony; Pedersen, Marius; Blahová, Jana; Hardeberg, Jon Yngve

    2015-01-01

    We introduce a new image database dedicated to multi-/hyperspectral image quality assessment. A total of nine scenes representing pseudo-at surfaces of different materials (textile, wood, skin. . . ) were captured by means of a 160 band hyperspectral system with a spectral range between 410 and 1000nm. Five spectral distortions were designed, applied to the spectral images and subsequently compared in a psychometric experiment, in order to provide a basis for applications such as the evaluation of spectral image difference measures. The database can be downloaded freely from http://www.colourlab.no/cid.

  18. Effects of task and image properties on visual-attention deployment in image-quality assessment

    NASA Astrophysics Data System (ADS)

    Alers, Hani; Redi, Judith; Liu, Hantao; Heynderickx, Ingrid

    2015-03-01

    It is important to understand how humans view images and how their behavior is affected by changes in the properties of the viewed images and the task they are given, particularly the task of scoring the image quality (IQ). This is a complex behavior that holds great importance for the field of image-quality research. This work builds upon 4 years of research work spanning three databases studying image-viewing behavior. Using eye-tracking equipment, it was possible to collect information on human viewing behavior of different kinds of stimuli and under different experimental settings. This work performs a cross-analysis on the results from all these databases using state-of-the-art similarity measures. The results strongly show that asking the viewers to score the IQ significantly changes their viewing behavior. Also muting the color saturation seems to affect the saliency of the images. However, a change in IQ was not consistently found to modify visual attention deployment, neither under free looking nor during scoring. These results are helpful in gaining a better understanding of image viewing behavior under different conditions. They also have important implications on work that collects subjective image-quality scores from human observers.

  19. Objective Quality Assessment for Color-to-Gray Image Conversion.

    PubMed

    Ma, Kede; Zhao, Tiesong; Zeng, Kai; Wang, Zhou

    2015-12-01

    Color-to-gray (C2G) image conversion is the process of transforming a color image into a grayscale one. Despite its wide usage in real-world applications, little work has been dedicated to compare the performance of C2G conversion algorithms. Subjective evaluation is reliable but is also inconvenient and time consuming. Here, we make one of the first attempts to develop an objective quality model that automatically predicts the perceived quality of C2G converted images. Inspired by the philosophy of the structural similarity index, we propose a C2G structural similarity (C2G-SSIM) index, which evaluates the luminance, contrast, and structure similarities between the reference color image and the C2G converted image. The three components are then combined depending on image type to yield an overall quality measure. Experimental results show that the proposed C2G-SSIM index has close agreement with subjective rankings and significantly outperforms existing objective quality metrics for C2G conversion. To explore the potentials of C2G-SSIM, we further demonstrate its use in two applications: 1) automatic parameter tuning for C2G conversion algorithms and 2) adaptive fusion of C2G converted images.

  20. MO-DE-209-03: Assessing Image Quality

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zhao, W.

    Digital Breast Tomosynthesis (DBT) is rapidly replacing mammography as the standard of care in breast cancer screening and diagnosis. DBT is a form of computed tomography, in which a limited set of projection images are acquired over a small angular range and reconstructed into tomographic data. The angular range varies from 15° to 50° and the number of projections varies between 9 and 25 projections, as determined by the equipment manufacturer. It is equally valid to treat DBT as the digital analog of classical tomography – that is, linear tomography. In fact, the name “tomosynthesis” stands for “synthetic tomography.” DBTmore » shares many common features with classical tomography, including the radiographic appearance, dose, and image quality considerations. As such, both the science and practical physics of DBT systems is a hybrid between computed tomography and classical tomographic methods. In this lecture, we will explore the continuum from radiography to computed tomography to illustrate the characteristics of DBT. This lecture will consist of four presentations that will provide a complete overview of DBT, including a review of the fundamentals of DBT acquisition, a discussion of DBT reconstruction methods, an overview of dosimetry for DBT systems, and summary of the underlying image theory of DBT thereby relating image quality and dose. Learning Objectives: To understand the fundamental principles behind tomosynthesis image acquisition. To understand the fundamentals of tomosynthesis image reconstruction. To learn the determinants of image quality and dose in DBT, including measurement techniques. To learn the image theory underlying tomosynthesis, and the relationship between dose and image quality. ADM is a consultant to, and holds stock in, Real Time Tomography, LLC. ADM receives research support from Hologic Inc., Analogic Inc., and Barco NV.; ADM is a member of the Scientific Advisory Board for Gamma Medica Inc.; A. Maidment, Research

  1. The effect of image quality, repeated study, and assessment method on anatomy learning.

    PubMed

    Fenesi, Barbara; Mackinnon, Chelsea; Cheng, Lucia; Kim, Joseph A; Wainman, Bruce C

    2017-06-01

    The use of two-dimensional (2D) images is consistently used to prepare anatomy students for handling real specimen. This study examined whether the quality of 2D images is a critical component in anatomy learning. The visual clarity and consistency of 2D anatomical images was systematically manipulated to produce low-quality and high-quality images of the human hand and human eye. On day 0, participants learned about each anatomical specimen from paper booklets using either low-quality or high-quality images, and then completed a comprehension test using either 2D images or three-dimensional (3D) cadaveric specimens. On day 1, participants relearned each booklet, and on day 2 participants completed a final comprehension test using either 2D images or 3D cadaveric specimens. The effect of image quality on learning varied according to anatomical content, with high-quality images having a greater effect on improving learning of hand anatomy than eye anatomy (high-quality vs. low-quality for hand anatomy P = 0.018; high-quality vs. low-quality for eye anatomy P = 0.247). Also, the benefit of high-quality images on hand anatomy learning was restricted to performance on short-answer (SA) questions immediately after learning (high-quality vs. low-quality on SA questions P = 0.018), but did not apply to performance on multiple-choice (MC) questions (high-quality vs. low-quality on MC questions P = 0.109) or after participants had an additional learning opportunity (24 hours later) with anatomy content (high vs. low on SA questions P = 0.643). This study underscores the limited impact of image quality on anatomy learning, and questions whether investment in enhancing image quality of learning aids significantly promotes knowledge development. Anat Sci Educ 10: 249-261. © 2016 American Association of Anatomists. © 2016 American Association of Anatomists.

  2. Age and Acceptance of Euthanasia.

    ERIC Educational Resources Information Center

    Ward, Russell A.

    1980-01-01

    Study explores relationship between age (and sex and race) and acceptance of euthanasia. Women and non-Whites were less accepting because of religiosity. Among older people less acceptance was attributable to their lesser education and greater religiosity. Results suggest that quality of life in old age affects acceptability of euthanasia. (Author)

  3. Multiple Image Arrangement for Subjective Quality Assessment

    NASA Astrophysics Data System (ADS)

    Wang, Yan; Zhai, Guangtao

    2017-12-01

    Subjective quality assessment serves as the foundation for almost all visual quality related researches. Size of the image quality databases has expanded from dozens to thousands in the last decades. Since each subjective rating therein has to be averaged over quite a few participants, the ever-increasing overall size of those databases calls for an evolution of existing subjective test methods. Traditional single/double stimulus based approaches are being replaced by multiple image tests, where several distorted versions of the original one are displayed and rated at once. And this naturally brings upon the question of how to arrange those multiple images on screen during the test. In this paper, we answer this question by performing subjective viewing test with eye tracker for different types arrangements. Our research indicates that isometric arrangement imposes less duress on participants and has more uniform distribution of eye fixations and movements and therefore is expected to generate more reliable subjective ratings.

  4. Correlations among steroid fear, acceptability, usage frequency, quality of life and disease severity in childhood eczema.

    PubMed

    Hon, Kam Lun; Tsang, Yin Ching K; Pong, Nga Hin; Luk, David C K; Lee, Vivian W; Woo, Wing Man; Lam, Chak Yiu Justin; Yeung, Yun Ting Eunice; Chau, Yiu Shing Sunny; Chui, Ka Kam Kenneth; Li, Ka Hin Gabriel; Leung, Ting Fan

    2015-10-01

    Topical corticosteroids (CSs) are the mainstay of treatment for eczema but CS phobia and fears are prevalent and influence therapeutic efficacy. To quantify if CS acceptability and fear affect patients' quality-of-life (QoL). Patients with eczema managed in the pediatric dermatology outpatient clinic of a university hospital were surveyed. Nottingham Eczema Severity Score (NESS) for severity, Children's Dermatology Life Quality Index (CDLQI) for QoL, CS fear, acceptability and reported frequency of CS use were measured with quantified questions. CS fears were prevalent among parents and caregivers of patients with eczema. Fifty-eight percent of parents reported general acceptability of CS as being very good or good, and many applied CS to their child regularly every week. However, >40% of parents reported CS fear "always" or "often", 41% reported that they "always" or "often" apply CS only when eczema got worse, 57% would discuss CS fear with their doctors, 30% would request CS-sparing medications and 14% "always" or "often" use traditional Chinese herbal medicine. Fears were predominantly interpersonal and less often iatrogenic in nature. Skin problems were the most concerned side effects of CS. CS acceptability, frequency of CS usage, CS fear and usage of alternative medications were independent domains in eczema management: CS fears correlated with CDLQI; CS usage frequency correlated with NESS and negatively with parental education; and CS acceptability correlated with parental education. Ordinal logistic regressions showed worse QoL was associated with more CS fear (odds ratio: 1.092 [95% CI: 1.023-1.165], p = 0.008). The extent of CS fears is independent of CS acceptability, but correlates with patients' QoL. Desensitization of parental CS fears should be integral part of eczema education and therapeutics in order to improve therapeutic efficacy and patients' QoL.

  5. Image Quality Analysis of Various Gastrointestinal Endoscopes: Why Image Quality Is a Prerequisite for Proper Diagnostic and Therapeutic Endoscopy

    PubMed Central

    Ko, Weon Jin; An, Pyeong; Ko, Kwang Hyun; Hahm, Ki Baik; Hong, Sung Pyo

    2015-01-01

    Arising from human curiosity in terms of the desire to look within the human body, endoscopy has undergone significant advances in modern medicine. Direct visualization of the gastrointestinal (GI) tract by traditional endoscopy was first introduced over 50 years ago, after which fairly rapid advancement from rigid esophagogastric scopes to flexible scopes and high definition videoscopes has occurred. In an effort towards early detection of precancerous lesions in the GI tract, several high-technology imaging scopes have been developed, including narrow band imaging, autofocus imaging, magnified endoscopy, and confocal microendoscopy. However, these modern developments have resulted in fundamental imaging technology being skewed towards red-green-blue and this technology has obscured the advantages of other endoscope techniques. In this review article, we have described the importance of image quality analysis using a survey to consider the diversity of endoscope system selection in order to better achieve diagnostic and therapeutic goals. The ultimate aims can be achieved through the adoption of modern endoscopy systems that obtain high image quality. PMID:26473119

  6. Technical Note: Unified imaging and robotic couch quality assurance.

    PubMed

    Cook, Molly C; Roper, Justin; Elder, Eric S; Schreibmann, Eduard

    2016-09-01

    To introduce a simplified quality assurance (QA) procedure that integrates tests for the linac's imaging components and the robotic couch. Current QA procedures for evaluating the alignment of the imaging system and linac require careful positioning of a phantom at isocenter before image acquisition and analysis. A complementary procedure for the robotic couch requires an initial displacement of the phantom and then evaluates the accuracy of repositioning the phantom at isocenter. We propose a two-in-one procedure that introduces a custom software module and incorporates both checks into one motion for increased efficiency. The phantom was manually set with random translational and rotational shifts, imaged with the in-room imaging system, and then registered to the isocenter using a custom software module. The software measured positioning accuracy by comparing the location of the repositioned phantom with a CAD model of the phantom at isocenter, which is physically verified using the MV port graticule. Repeatability of the custom software was tested by an assessment of internal marker location extraction on a series of scans taken over differing kV and CBCT acquisition parameters. The proposed method was able to correctly position the phantom at isocenter within acceptable 1 mm and 1° SRS tolerances, verified by both physical inspection and the custom software. Residual errors for mechanical accuracy were 0.26 mm vertically, 0.21 mm longitudinally, 0.55 mm laterally, 0.21° in pitch, 0.1° in roll, and 0.67° in yaw. The software module was shown to be robust across various scan acquisition parameters, detecting markers within 0.15 mm translationally in kV acquisitions and within 0.5 mm translationally and 0.3° rotationally across CBCT acquisitions with significant variations in voxel size. Agreement with vendor registration methods was well within 0.5 mm; differences were not statistically significant. As compared to the current two-step approach, the proposed

  7. Saliency image of feature building for image quality assessment

    NASA Astrophysics Data System (ADS)

    Ju, Xinuo; Sun, Jiyin; Wang, Peng

    2011-11-01

    The purpose and method of image quality assessment are quite different for automatic target recognition (ATR) and traditional application. Local invariant feature detectors, mainly including corner detectors, blob detectors and region detectors etc., are widely applied for ATR. A saliency model of feature was proposed to evaluate feasibility of ATR in this paper. The first step consisted of computing the first-order derivatives on horizontal orientation and vertical orientation, and computing DoG maps in different scales respectively. Next, saliency images of feature were built based auto-correlation matrix in different scale. Then, saliency images of feature of different scales amalgamated. Experiment were performed on a large test set, including infrared images and optical images, and the result showed that the salient regions computed by this model were consistent with real feature regions computed by mostly local invariant feature extraction algorithms.

  8. 21 CFR 820.86 - Acceptance status.

    Code of Federal Regulations, 2010 CFR

    2010-04-01

    ... QUALITY SYSTEM REGULATION Acceptance Activities § 820.86 Acceptance status. Each manufacturer shall... product which has passed the required acceptance activities is distributed, used, or installed. ...

  9. SU-E-P-11: Comparison of Image Quality and Radiation Dose Between Different Scanner System in Routine Abdomen CT

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Liao, S; Wang, Y; Weng, H

    Purpose To evaluate image quality and radiation dose of routine abdomen computed tomography exam with the automatic current modulation technique (ATCM) performed in two different brand 64-slice CT scanners in our site. Materials and Methods A retrospective review of routine abdomen CT exam performed with two scanners; scanner A and scanner B in our site. To calculate standard deviation of the portal hepatic level with a region of interest of 12.5 mm x 12.5mm represented to the image noise. The radiation dose was obtained from CT DICOM image information. Using Computed tomography dose index volume (CTDIv) to represented CT radiationmore » dose. The patient data in this study were with normal weight (about 65–75 Kg). Results The standard deviation of Scanner A was smaller than scanner B, the scanner A might with better image quality than scanner B. On the other hand, the radiation dose of scanner A was higher than scanner B(about higher 50–60%) with ATCM. Both of them, the radiation dose was under diagnostic reference level. Conclusion The ATCM systems in modern CT scanners can contribute a significant reduction in radiation dose to the patient. But the reduction by ATCM systems from different CT scanner manufacturers has slightly variation. Whatever CT scanner we use, it is necessary to find the acceptable threshold of image quality with the minimum possible radiation exposure to the patient in agreement with the ALARA principle.« less

  10. Image quality (IQ) guided multispectral image compression

    NASA Astrophysics Data System (ADS)

    Zheng, Yufeng; Chen, Genshe; Wang, Zhonghai; Blasch, Erik

    2016-05-01

    Image compression is necessary for data transportation, which saves both transferring time and storage space. In this paper, we focus on our discussion on lossy compression. There are many standard image formats and corresponding compression algorithms, for examples, JPEG (DCT -- discrete cosine transform), JPEG 2000 (DWT -- discrete wavelet transform), BPG (better portable graphics) and TIFF (LZW -- Lempel-Ziv-Welch). The image quality (IQ) of decompressed image will be measured by numerical metrics such as root mean square error (RMSE), peak signal-to-noise ratio (PSNR), and structural Similarity (SSIM) Index. Given an image and a specified IQ, we will investigate how to select a compression method and its parameters to achieve an expected compression. Our scenario consists of 3 steps. The first step is to compress a set of interested images by varying parameters and compute their IQs for each compression method. The second step is to create several regression models per compression method after analyzing the IQ-measurement versus compression-parameter from a number of compressed images. The third step is to compress the given image with the specified IQ using the selected compression method (JPEG, JPEG2000, BPG, or TIFF) according to the regressed models. The IQ may be specified by a compression ratio (e.g., 100), then we will select the compression method of the highest IQ (SSIM, or PSNR). Or the IQ may be specified by a IQ metric (e.g., SSIM = 0.8, or PSNR = 50), then we will select the compression method of the highest compression ratio. Our experiments tested on thermal (long-wave infrared) images (in gray scales) showed very promising results.

  11. Image quality of conventional images of dual-layer SPECTRAL CT: A phantom study.

    PubMed

    van Ommen, Fasco; Bennink, Edwin; Vlassenbroek, Alain; Dankbaar, Jan Willem; Schilham, Arnold M R; Viergever, Max A; de Jong, Hugo W A M

    2018-05-10

    Spectral CT using a dual layer detector offers the possibility of retrospectively introducing spectral information to conventional CT images. In theory, the dual-layer technology should not come with a dose or image quality penalty for conventional images. In this study, we evaluate the influence of a dual-layer detector (IQon Spectral CT, Philips Healthcare) on the image quality of conventional CT images, by comparing these images with those of a conventional but otherwise technically comparable single-layer CT scanner (Brilliance iCT, Philips Healthcare), by means of phantom experiments. For both CT scanners, conventional CT images were acquired using four adult scanning protocols: (a) body helical, (b) body axial, (c) head helical, and (d) head axial. A CATPHAN 600 phantom was scanned to conduct an assessment of image quality metrics at equivalent (CTDI) dose levels. Noise was characterized by means of noise power spectra (NPS) and standard deviation (SD) of a uniform region, and spatial resolution was evaluated with modulation transfer functions (MTF) of a tungsten wire. In addition, contrast-to-noise ratio (CNR), image uniformity, CT number linearity, slice thickness, slice spacing, and spatial linearity were measured and evaluated. Additional measurements of CNR, resolution and noise were performed in two larger phantoms. The resolution levels at 50%, 10%, and 5% MTF of the iCT and IQon showed small, but significant differences up to 0.25 lp/cm for body scans, and up to 0.2 lp/cm for head scans in favor of the IQon. The iCT and IQon showed perfect CT linearity for body scans, but for head scans both scanners showed an underestimation of the CT numbers of materials with a high opacity. Slice thickness was slightly overestimated for both scanners. Slice spacing was comparable and reconstructed correctly. In addition, spatial linearity was excellent for both scanners, with a maximum error of 0.11 mm. CNR was higher on the IQon compared to the iCT for both normal

  12. Fuzzy intelligent quality monitoring model for X-ray image processing.

    PubMed

    Khalatbari, Azadeh; Jenab, Kouroush

    2009-01-01

    Today's imaging diagnosis needs to adapt modern techniques of quality engineering to maintain and improve its accuracy and reliability in health care system. One of the main factors that influences diagnostic accuracy of plain film X-ray on detecting pathology is the level of film exposure. If the level of film exposure is not adequate, a normal body structure may be interpretated as pathology and vice versa. This not only influences the patient management but also has an impact on health care cost and patient's quality of life. Therefore, providing an accurate and high quality image is the first step toward an excellent patient management in any health care system. In this paper, we study these techniques and also present a fuzzy intelligent quality monitoring model, which can be used to keep variables from degrading the image quality. The variables derived from chemical activity, cleaning procedures, maintenance, and monitoring may not be sensed, measured, or calculated precisely due to uncertain situations. Therefore, the gamma-level fuzzy Bayesian model for quality monitoring of an image processing is proposed. In order to apply the Bayesian concept, the fuzzy quality characteristics are assumed as fuzzy random variables. Using the fuzzy quality characteristics, the newly developed model calculates the degradation risk for image processing. A numerical example is also presented to demonstrate the application of the model.

  13. Backward Registration Based Aspect Ratio Similarity (ARS) for Image Retargeting Quality Assessment.

    PubMed

    Zhang, Yabin; Fang, Yuming; Lin, Weisi; Zhang, Xinfeng; Li, Leida

    2016-06-28

    During the past few years, there have been various kinds of content-aware image retargeting operators proposed for image resizing. However, the lack of effective objective retargeting quality assessment metrics limits the further development of image retargeting techniques. Different from traditional Image Quality Assessment (IQA) metrics, the quality degradation during image retargeting is caused by artificial retargeting modifications, and the difficulty for Image Retargeting Quality Assessment (IRQA) lies in the alternation of the image resolution and content, which makes it impossible to directly evaluate the quality degradation like traditional IQA. In this paper, we interpret the image retargeting in a unified framework of resampling grid generation and forward resampling. We show that the geometric change estimation is an efficient way to clarify the relationship between the images. We formulate the geometric change estimation as a Backward Registration problem with Markov Random Field (MRF) and provide an effective solution. The geometric change aims to provide the evidence about how the original image is resized into the target image. Under the guidance of the geometric change, we develop a novel Aspect Ratio Similarity metric (ARS) to evaluate the visual quality of retargeted images by exploiting the local block changes with a visual importance pooling strategy. Experimental results on the publicly available MIT RetargetMe and CUHK datasets demonstrate that the proposed ARS can predict more accurate visual quality of retargeted images compared with state-of-the-art IRQA metrics.

  14. Image quality testing of assembled IR camera modules

    NASA Astrophysics Data System (ADS)

    Winters, Daniel; Erichsen, Patrik

    2013-10-01

    Infrared (IR) camera modules for the LWIR (8-12_m) that combine IR imaging optics with microbolometer focal plane array (FPA) sensors with readout electronics are becoming more and more a mass market product. At the same time, steady improvements in sensor resolution in the higher priced markets raise the requirement for imaging performance of objectives and the proper alignment between objective and FPA. This puts pressure on camera manufacturers and system integrators to assess the image quality of finished camera modules in a cost-efficient and automated way for quality control or during end-of-line testing. In this paper we present recent development work done in the field of image quality testing of IR camera modules. This technology provides a wealth of additional information in contrast to the more traditional test methods like minimum resolvable temperature difference (MRTD) which give only a subjective overall test result. Parameters that can be measured are image quality via the modulation transfer function (MTF) for broadband or with various bandpass filters on- and off-axis and optical parameters like e.g. effective focal length (EFL) and distortion. If the camera module allows for refocusing the optics, additional parameters like best focus plane, image plane tilt, auto-focus quality, chief ray angle etc. can be characterized. Additionally, the homogeneity and response of the sensor with the optics can be characterized in order to calculate the appropriate tables for non-uniformity correction (NUC). The technology can also be used to control active alignment methods during mechanical assembly of optics to high resolution sensors. Other important points that are discussed are the flexibility of the technology to test IR modules with different form factors, electrical interfaces and last but not least the suitability for fully automated measurements in mass production.

  15. An approach for quantitative image quality analysis for CT

    NASA Astrophysics Data System (ADS)

    Rahimi, Amir; Cochran, Joe; Mooney, Doug; Regensburger, Joe

    2016-03-01

    An objective and standardized approach to assess image quality of Compute Tomography (CT) systems is required in a wide variety of imaging processes to identify CT systems appropriate for a given application. We present an overview of the framework we have developed to help standardize and to objectively assess CT image quality for different models of CT scanners used for security applications. Within this framework, we have developed methods to quantitatively measure metrics that should correlate with feature identification, detection accuracy and precision, and image registration capabilities of CT machines and to identify strengths and weaknesses in different CT imaging technologies in transportation security. To that end we have designed, developed and constructed phantoms that allow for systematic and repeatable measurements of roughly 88 image quality metrics, representing modulation transfer function, noise equivalent quanta, noise power spectra, slice sensitivity profiles, streak artifacts, CT number uniformity, CT number consistency, object length accuracy, CT number path length consistency, and object registration. Furthermore, we have developed a sophisticated MATLAB based image analysis tool kit to analyze CT generated images of phantoms and report these metrics in a format that is standardized across the considered models of CT scanners, allowing for comparative image quality analysis within a CT model or between different CT models. In addition, we have developed a modified sparse principal component analysis (SPCA) method to generate a modified set of PCA components as compared to the standard principal component analysis (PCA) with sparse loadings in conjunction with Hotelling T2 statistical analysis method to compare, qualify, and detect faults in the tested systems.

  16. Automated characterization of perceptual quality of clinical chest radiographs: Validation and calibration to observer preference

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Samei, Ehsan, E-mail: samei@duke.edu; Lin, Yuan; Choudhury, Kingshuk R.

    Purpose: The authors previously proposed an image-based technique [Y. Lin et al. Med. Phys. 39, 7019–7031 (2012)] to assess the perceptual quality of clinical chest radiographs. In this study, an observer study was designed and conducted to validate the output of the program against rankings by expert radiologists and to establish the ranges of the output values that reflect the acceptable image appearance so the program output can be used for image quality optimization and tracking. Methods: Using an IRB-approved protocol, 2500 clinical chest radiographs (PA/AP) were collected from our clinical operation. The images were processed through our perceptual qualitymore » assessment program to measure their appearance in terms of ten metrics of perceptual image quality: lung gray level, lung detail, lung noise, rib–lung contrast, rib sharpness, mediastinum detail, mediastinum noise, mediastinum alignment, subdiaphragm–lung contrast, and subdiaphragm area. From the results, for each targeted appearance attribute/metric, 18 images were selected such that the images presented a relatively constant appearance with respect to all metrics except the targeted one. The images were then incorporated into a graphical user interface, which displayed them into three panels of six in a random order. Using a DICOM calibrated diagnostic display workstation and under low ambient lighting conditions, each of five participating attending chest radiologists was tasked to spatially order the images based only on the targeted appearance attribute regardless of the other qualities. Once ordered, the observer also indicated the range of image appearances that he/she considered clinically acceptable. The observer data were analyzed in terms of the correlations between the observer and algorithmic rankings and interobserver variability. An observer-averaged acceptable image appearance was also statistically derived for each quality attribute based on the collected individual acceptable

  17. Correlation of the clinical and physical image quality in chest radiography for average adults with a computed radiography imaging system.

    PubMed

    Moore, C S; Wood, T J; Beavis, A W; Saunderson, J R

    2013-07-01

    The purpose of this study was to examine the correlation between the quality of visually graded patient (clinical) chest images and a quantitative assessment of chest phantom (physical) images acquired with a computed radiography (CR) imaging system. The results of a previously published study, in which four experienced image evaluators graded computer-simulated postero-anterior chest images using a visual grading analysis scoring (VGAS) scheme, were used for the clinical image quality measurement. Contrast-to-noise ratio (CNR) and effective dose efficiency (eDE) were used as physical image quality metrics measured in a uniform chest phantom. Although optimal values of these physical metrics for chest radiography were not derived in this work, their correlation with VGAS in images acquired without an antiscatter grid across the diagnostic range of X-ray tube voltages was determined using Pearson's correlation coefficient. Clinical and physical image quality metrics increased with decreasing tube voltage. Statistically significant correlations between VGAS and CNR (R=0.87, p<0.033) and eDE (R=0.77, p<0.008) were observed. Medical physics experts may use the physical image quality metrics described here in quality assurance programmes and optimisation studies with a degree of confidence that they reflect the clinical image quality in chest CR images acquired without an antiscatter grid. A statistically significant correlation has been found between the clinical and physical image quality in CR chest imaging. The results support the value of using CNR and eDE in the evaluation of quality in clinical thorax radiography.

  18. Correlation of the clinical and physical image quality in chest radiography for average adults with a computed radiography imaging system

    PubMed Central

    Wood, T J; Beavis, A W; Saunderson, J R

    2013-01-01

    Objective: The purpose of this study was to examine the correlation between the quality of visually graded patient (clinical) chest images and a quantitative assessment of chest phantom (physical) images acquired with a computed radiography (CR) imaging system. Methods: The results of a previously published study, in which four experienced image evaluators graded computer-simulated postero-anterior chest images using a visual grading analysis scoring (VGAS) scheme, were used for the clinical image quality measurement. Contrast-to-noise ratio (CNR) and effective dose efficiency (eDE) were used as physical image quality metrics measured in a uniform chest phantom. Although optimal values of these physical metrics for chest radiography were not derived in this work, their correlation with VGAS in images acquired without an antiscatter grid across the diagnostic range of X-ray tube voltages was determined using Pearson’s correlation coefficient. Results: Clinical and physical image quality metrics increased with decreasing tube voltage. Statistically significant correlations between VGAS and CNR (R=0.87, p<0.033) and eDE (R=0.77, p<0.008) were observed. Conclusion: Medical physics experts may use the physical image quality metrics described here in quality assurance programmes and optimisation studies with a degree of confidence that they reflect the clinical image quality in chest CR images acquired without an antiscatter grid. Advances in knowledge: A statistically significant correlation has been found between the clinical and physical image quality in CR chest imaging. The results support the value of using CNR and eDE in the evaluation of quality in clinical thorax radiography. PMID:23568362

  19. Dose reduction in abdominal computed tomography: intraindividual comparison of image quality of full-dose standard and half-dose iterative reconstructions with dual-source computed tomography.

    PubMed

    May, Matthias S; Wüst, Wolfgang; Brand, Michael; Stahl, Christian; Allmendinger, Thomas; Schmidt, Bernhard; Uder, Michael; Lell, Michael M

    2011-07-01

    We sought to evaluate the image quality of iterative reconstruction in image space (IRIS) in half-dose (HD) datasets compared with full-dose (FD) and HD filtered back projection (FBP) reconstruction in abdominal computed tomography (CT). To acquire data with FD and HD simultaneously, contrast-enhanced abdominal CT was performed with a dual-source CT system, both tubes operating at 120 kV, 100 ref.mAs, and pitch 0.8. Three different image datasets were reconstructed from the raw data: Standard FD images applying FBP which served as reference, HD images applying FBP and HD images applying IRIS. For the HD data sets, only data from 1 tube detector-system was used. Quantitative image quality analysis was performed by measuring image noise in tissue and air. Qualitative image quality was evaluated according to the European Guidelines on Quality criteria for CT. Additional assessment of artifacts, lesion conspicuity, and edge sharpness was performed. : Image noise in soft tissue was substantially decreased in HD-IRIS (-3.4 HU, -22%) and increased in HD-FBP (+6.2 HU, +39%) images when compared with the reference (mean noise, 15.9 HU). No significant differences between the FD-FBP and HD-IRIS images were found for the visually sharp anatomic reproduction, overall diagnostic acceptability (P = 0.923), lesion conspicuity (P = 0.592), and edge sharpness (P = 0.589), while HD-FBP was rated inferior. Streak artifacts and beam hardening was significantly more prominent in HD-FBP while HD-IRIS images exhibited a slightly different noise pattern. Direct intrapatient comparison of standard FD body protocols and HD-IRIS reconstruction suggest that the latest iterative reconstruction algorithms allow for approximately 50% dose reduction without deterioration of the high image quality necessary for confident diagnosis.

  20. Real-time computer treatment of THz passive device images with the high image quality

    NASA Astrophysics Data System (ADS)

    Trofimov, Vyacheslav A.; Trofimov, Vladislav V.

    2012-06-01

    We demonstrate real-time computer code improving significantly the quality of images captured by the passive THz imaging system. The code is not only designed for a THz passive device: it can be applied to any kind of such devices and active THz imaging systems as well. We applied our code for computer processing of images captured by four passive THz imaging devices manufactured by different companies. It should be stressed that computer processing of images produced by different companies requires using the different spatial filters usually. The performance of current version of the computer code is greater than one image per second for a THz image having more than 5000 pixels and 24 bit number representation. Processing of THz single image produces about 20 images simultaneously corresponding to various spatial filters. The computer code allows increasing the number of pixels for processed images without noticeable reduction of image quality. The performance of the computer code can be increased many times using parallel algorithms for processing the image. We develop original spatial filters which allow one to see objects with sizes less than 2 cm. The imagery is produced by passive THz imaging devices which captured the images of objects hidden under opaque clothes. For images with high noise we develop an approach which results in suppression of the noise after using the computer processing and we obtain the good quality image. With the aim of illustrating the efficiency of the developed approach we demonstrate the detection of the liquid explosive, ordinary explosive, knife, pistol, metal plate, CD, ceramics, chocolate and other objects hidden under opaque clothes. The results demonstrate the high efficiency of our approach for the detection of hidden objects and they are a very promising solution for the security problem.

  1. "It's all about acceptance": A qualitative study exploring a model of positive body image for people with spinal cord injury.

    PubMed

    Bailey, K Alysse; Gammage, Kimberley L; van Ingen, Cathy; Ditor, David S

    2015-09-01

    Using modified constructivist grounded theory, the purpose of the present study was to explore positive body image experiences in people with spinal cord injury. Nine participants (five women, four men) varying in age (21-63 years), type of injury (C3-T7; complete and incomplete), and years post-injury (4-36 years) were recruited. The following main categories were found: body acceptance, body appreciation and gratitude, social support, functional gains, independence, media literacy, broadly conceptualizing beauty, inner positivity influencing outer demeanour, finding others who have a positive body image, unconditional acceptance from others, religion/spirituality, listening to and taking care of the body, managing secondary complications, minimizing pain, and respect. Interestingly, there was consistency in positive body image characteristics reported in this study with those found in previous research, demonstrating universality of positive body image. However, unique characteristics (e.g., resilience, functional gains, independence) were also reported demonstrating the importance of exploring positive body image in diverse groups. Copyright © 2015 Elsevier Ltd. All rights reserved.

  2. The mobile image quality survey game

    NASA Astrophysics Data System (ADS)

    Rasmussen, D. René

    2012-01-01

    In this paper we discuss human assessment of the quality of photographic still images, that are degraded in various manners relative to an original, for example due to compression or noise. In particular, we examine and present results from a technique where observers view images on a mobile device, perform pairwise comparisons, identify defects in the images, and interact with the display to indicate the location of the defects. The technique measures the response time and accuracy of the responses. By posing the survey in a form similar to a game, providing performance feedback to the observer, the technique attempts to increase the engagement of the observers, and to avoid exhausting observers, a factor that is often a problem for subjective surveys. The results are compared with the known physical magnitudes of the defects and with results from similar web-based surveys. The strengths and weaknesses of the technique are discussed. Possible extensions of the technique to video quality assessment are also discussed.

  3. Perceived assessment metrics for visible and infrared color fused image quality without reference image

    NASA Astrophysics Data System (ADS)

    Yu, Xuelian; Chen, Qian; Gu, Guohua; Ren, Jianle; Sui, Xiubao

    2015-02-01

    Designing objective quality assessment of color-fused image is a very demanding and challenging task. We propose four no-reference metrics based on human visual system characteristics for objectively evaluating the quality of false color fusion image. The perceived edge metric (PEM) is defined based on visual perception model and color image gradient similarity between the fused image and the source images. The perceptual contrast metric (PCM) is established associating multi-scale contrast and varying contrast sensitivity filter (CSF) with color components. The linear combination of the standard deviation and mean value over the fused image construct the image colorfulness metric (ICM). The color comfort metric (CCM) is designed by the average saturation and the ratio of pixels with high and low saturation. The qualitative and quantitative experimental results demonstrate that the proposed metrics have a good agreement with subjective perception.

  4. SU-E-I-43: Pediatric CT Dose and Image Quality Optimization

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Stevens, G; Singh, R

    2014-06-01

    Purpose: To design an approach to optimize radiation dose and image quality for pediatric CT imaging, and to evaluate expected performance. Methods: A methodology was designed to quantify relative image quality as a function of CT image acquisition parameters. Image contrast and image noise were used to indicate expected conspicuity of objects, and a wide-cone system was used to minimize scan time for motion avoidance. A decision framework was designed to select acquisition parameters as a weighted combination of image quality and dose. Phantom tests were used to acquire images at multiple techniques to demonstrate expected contrast, noise and dose.more » Anthropomorphic phantoms with contrast inserts were imaged on a 160mm CT system with tube voltage capabilities as low as 70kVp. Previously acquired clinical images were used in conjunction with simulation tools to emulate images at different tube voltages and currents to assess human observer preferences. Results: Examination of image contrast, noise, dose and tube/generator capabilities indicates a clinical task and object-size dependent optimization. Phantom experiments confirm that system modeling can be used to achieve the desired image quality and noise performance. Observer studies indicate that clinical utilization of this optimization requires a modified approach to achieve the desired performance. Conclusion: This work indicates the potential to optimize radiation dose and image quality for pediatric CT imaging. In addition, the methodology can be used in an automated parameter selection feature that can suggest techniques given a limited number of user inputs. G Stevens and R Singh are employees of GE Healthcare.« less

  5. Measuring saliency in images: which experimental parameters for the assessment of image quality?

    NASA Astrophysics Data System (ADS)

    Fredembach, Clement; Woolfe, Geoff; Wang, Jue

    2012-01-01

    Predicting which areas of an image are perceptually salient or attended to has become an essential pre-requisite of many computer vision applications. Because observers are notoriously unreliable in remembering where they look a posteriori, and because asking where they look while observing the image necessarily in uences the results, ground truth about saliency and visual attention has to be obtained by gaze tracking methods. From the early work of Buswell and Yarbus to the most recent forays in computer vision there has been, perhaps unfortunately, little agreement on standardisation of eye tracking protocols for measuring visual attention. As the number of parameters involved in experimental methodology can be large, their individual in uence on the nal results is not well understood. Consequently, the performance of saliency algorithms, when assessed by correlation techniques, varies greatly across the literature. In this paper, we concern ourselves with the problem of image quality. Specically: where people look when judging images. We show that in this case, the performance gap between existing saliency prediction algorithms and experimental results is signicantly larger than otherwise reported. To understand this discrepancy, we rst devise an experimental protocol that is adapted to the task of measuring image quality. In a second step, we compare our experimental parameters with the ones of existing methods and show that a lot of the variability can directly be ascribed to these dierences in experimental methodology and choice of variables. In particular, the choice of a task, e.g., judging image quality vs. free viewing, has a great impact on measured saliency maps, suggesting that even for a mildly cognitive task, ground truth obtained by free viewing does not adapt well. Careful analysis of the prior art also reveals that systematic bias can occur depending on instrumental calibration and the choice of test images. We conclude this work by proposing a

  6. Boosting Quality Registries with Clinical Decision Support Functionality*. User Acceptance of a Prototype Applied to HIV/TB Drug Therapy.

    PubMed

    Wannheden, Carolina; Hvitfeldt-Forsberg, Helena; Eftimovska, Elena; Westling, Katarina; Ellenius, Johan

    2017-08-11

    The care of HIV-related tuberculosis (HIV/TB) is complex and challenging. Clinical decision support (CDS) systems can contribute to improve quality of care, but more knowledge is needed on factors determining user acceptance of CDS. To analyze physicians' and nurses' acceptance of a CDS prototype for evidence-based drug therapy recommendations for HIV/TB treatment. Physicians and nurses were involved in designing a CDS prototype intended for future integration with the Swedish national HIV quality registry. Focus group evaluation was performed with ten nurses and four physicians, respectively. The Unified Theory of Acceptance and Use of Technology (UTAUT) was used to analyze acceptance. We identified several potential benefits with the CDS prototype as well as some concerns that could be addressed by redesign. There was also concern about dependence on physician attitudes, as well as technical, organizational, and legal issues. Acceptance evaluation at a prototype stage provided rich data to improve the future design of a CDS prototype. Apart from design and development efforts, substantial organizational efforts are needed to enable the implementation and maintenance of a future CDS system.

  7. A Perceptually Weighted Rank Correlation Indicator for Objective Image Quality Assessment

    NASA Astrophysics Data System (ADS)

    Wu, Qingbo; Li, Hongliang; Meng, Fanman; Ngan, King N.

    2018-05-01

    In the field of objective image quality assessment (IQA), the Spearman's $\\rho$ and Kendall's $\\tau$ are two most popular rank correlation indicators, which straightforwardly assign uniform weight to all quality levels and assume each pair of images are sortable. They are successful for measuring the average accuracy of an IQA metric in ranking multiple processed images. However, two important perceptual properties are ignored by them as well. Firstly, the sorting accuracy (SA) of high quality images are usually more important than the poor quality ones in many real world applications, where only the top-ranked images would be pushed to the users. Secondly, due to the subjective uncertainty in making judgement, two perceptually similar images are usually hardly sortable, whose ranks do not contribute to the evaluation of an IQA metric. To more accurately compare different IQA algorithms, we explore a perceptually weighted rank correlation indicator in this paper, which rewards the capability of correctly ranking high quality images, and suppresses the attention towards insensitive rank mistakes. More specifically, we focus on activating `valid' pairwise comparison towards image quality, whose difference exceeds a given sensory threshold (ST). Meanwhile, each image pair is assigned an unique weight, which is determined by both the quality level and rank deviation. By modifying the perception threshold, we can illustrate the sorting accuracy with a more sophisticated SA-ST curve, rather than a single rank correlation coefficient. The proposed indicator offers a new insight for interpreting visual perception behaviors. Furthermore, the applicability of our indicator is validated in recommending robust IQA metrics for both the degraded and enhanced image data.

  8. Spectral CT imaging in patients with Budd-Chiari syndrome: investigation of image quality.

    PubMed

    Su, Lei; Dong, Junqiang; Sun, Qiang; Liu, Jie; Lv, Peijie; Hu, Lili; Yan, Liangliang; Gao, Jianbo

    2014-11-01

    To assess the image quality of monochromatic imaging from spectral CT in patients with Budd-Chiari syndrome (BCS), fifty patients with BCS underwent spectral CT to generate conventional 140 kVp polychromatic images (group A) and monochromatic images, with energy levels from 40 to 80, 40 + 70, and 50 + 70 keV fusion images (group B) during the portal venous phase (PVP) and the hepatic venous phase (HVP). Two-sample t tests compared vessel-to-liver contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) for the portal vein (PV), hepatic vein (HV), inferior vena cava. Readers' subjective evaluations of the image quality were recorded. The highest SNR values in group B were distributed at 50 keV; the highest CNR values in group B were distributed at 40 keV. The higher CNR values and SNR values were obtained though PVP of PV (SNR 18.39 ± 6.13 vs. 10.56 ± 3.31, CNR 7.81 ± 3.40 vs. 3.58 ± 1.31) and HVP of HV (3.89 ± 2.08 vs. 1.27 ± 1.55) in the group B; the lower image noise for group B was at 70 keV and 50 + 70 keV (15.54 ± 8.39 vs. 18.40 ± 4.97, P = 0.0004 and 18.97 ± 7.61 vs. 18.40 ± 4.97, P = 0.0691); the results show that the 50 + 70 keV fusion image quality was better than that in group A. Monochromatic energy levels of 40-70, 40 + 70, and 50 + 70 keV fusion image can increase vascular contrast and that will be helpful for the diagnosis of BCS, we select the 50 + 70 keV fusion image to acquire the best BCS images.

  9. Reducing radiation dose without compromising image quality in preoperative perforator flap imaging with CTA using ASIR technology.

    PubMed

    Niumsawatt, Vachara; Debrotwir, Andrew N; Rozen, Warren Matthew

    2014-01-01

    Computed tomographic angiography (CTA) has become a mainstay in preoperative perforator flap planning in the modern era of reconstructive surgery. However, the increased use of CTA does raise the concern of radiation exposure to patients. Several techniques have been developed to decrease radiation dosage without compromising image quality, with varying results. The most recent advance is in the improvement of image reconstruction using an adaptive statistical iterative reconstruction (ASIR) algorithm. We sought to evaluate the image quality of ASIR in preoperative deep inferior epigastric perforator (DIEP) flap surgery, through a direct comparison with conventional filtered back projection (FBP) images. A prospective review of 60 consecutive ASIR and 60 consecutive FBP CTA images using similar protocol (except for radiation dosage) was undertaken, analyzed by 2 independent reviewers. In both groups, we were able to accurately identify axial arteries and their perforators. Subjective analysis of image quality demonstrated no statistically significant difference between techniques. ASIR can thus be used for preoperative imaging with similar image quality to FBP, but with a 60% reduction in radiation delivery to patients.

  10. Dosimetry and image quality assessment in a direct radiography system

    PubMed Central

    Oliveira, Bruno Beraldo; de Oliveira, Marcio Alves; Paixão, Lucas; Teixeira, Maria Helena Araújo; Nogueira, Maria do Socorro

    2014-01-01

    Objective To evaluate the mean glandular dose with a solid state detector and the image quality in a direct radiography system, utilizing phantoms. Materials and Methods Irradiations were performed with automatic exposure control and polymethyl methacrylate slabs with different thicknesses to calculate glandular dose values. The image quality was evaluated by means of the structures visualized on the images of the phantoms. Results Considering the uncertainty of the measurements, the mean glandular dose results are in agreement with the values provided by the equipment and with internationally adopted reference levels. Results obtained from images of the phantoms were in agreement with the reference values. Conclusion The present study contributes to verify the equipment conformity as regards dose values and image quality. PMID:25741119

  11. Can image enhancement allow radiation dose to be reduced whilst maintaining the perceived diagnostic image quality required for coronary angiography?

    PubMed Central

    Joshi, Anuja; Gislason-Lee, Amber J; Keeble, Claire; Sivananthan, Uduvil M

    2017-01-01

    Objective: The aim of this research was to quantify the reduction in radiation dose facilitated by image processing alone for percutaneous coronary intervention (PCI) patient angiograms, without reducing the perceived image quality required to confidently make a diagnosis. Methods: Incremental amounts of image noise were added to five PCI angiograms, simulating the angiogram as having been acquired at corresponding lower dose levels (10–89% dose reduction). 16 observers with relevant experience scored the image quality of these angiograms in 3 states—with no image processing and with 2 different modern image processing algorithms applied. These algorithms are used on state-of-the-art and previous generation cardiac interventional X-ray systems. Ordinal regression allowing for random effects and the delta method were used to quantify the dose reduction possible by the processing algorithms, for equivalent image quality scores. Results: Observers rated the quality of the images processed with the state-of-the-art and previous generation image processing with a 24.9% and 15.6% dose reduction, respectively, as equivalent in quality to the unenhanced images. The dose reduction facilitated by the state-of-the-art image processing relative to previous generation processing was 10.3%. Conclusion: Results demonstrate that statistically significant dose reduction can be facilitated with no loss in perceived image quality using modern image enhancement; the most recent processing algorithm was more effective in preserving image quality at lower doses. Advances in knowledge: Image enhancement was shown to maintain perceived image quality in coronary angiography at a reduced level of radiation dose using computer software to produce synthetic images from real angiograms simulating a reduction in dose. PMID:28124572

  12. Identification of suitable fundus images using automated quality assessment methods.

    PubMed

    Şevik, Uğur; Köse, Cemal; Berber, Tolga; Erdöl, Hidayet

    2014-04-01

    Retinal image quality assessment (IQA) is a crucial process for automated retinal image analysis systems to obtain an accurate and successful diagnosis of retinal diseases. Consequently, the first step in a good retinal image analysis system is measuring the quality of the input image. We present an approach for finding medically suitable retinal images for retinal diagnosis. We used a three-class grading system that consists of good, bad, and outlier classes. We created a retinal image quality dataset with a total of 216 consecutive images called the Diabetic Retinopathy Image Database. We identified the suitable images within the good images for automatic retinal image analysis systems using a novel method. Subsequently, we evaluated our retinal image suitability approach using the Digital Retinal Images for Vessel Extraction and Standard Diabetic Retinopathy Database Calibration level 1 public datasets. The results were measured through the F1 metric, which is a harmonic mean of precision and recall metrics. The highest F1 scores of the IQA tests were 99.60%, 96.50%, and 85.00% for good, bad, and outlier classes, respectively. Additionally, the accuracy of our suitable image detection approach was 98.08%. Our approach can be integrated into any automatic retinal analysis system with sufficient performance scores.

  13. Automated daily quality control analysis for mammography in a multi-unit imaging center.

    PubMed

    Sundell, Veli-Matti; Mäkelä, Teemu; Meaney, Alexander; Kaasalainen, Touko; Savolainen, Sauli

    2018-01-01

    Background The high requirements for mammography image quality necessitate a systematic quality assurance process. Digital imaging allows automation of the image quality analysis, which can potentially improve repeatability and objectivity compared to a visual evaluation made by the users. Purpose To develop an automatic image quality analysis software for daily mammography quality control in a multi-unit imaging center. Material and Methods An automated image quality analysis software using the discrete wavelet transform and multiresolution analysis was developed for the American College of Radiology accreditation phantom. The software was validated by analyzing 60 randomly selected phantom images from six mammography systems and 20 phantom images with different dose levels from one mammography system. The results were compared to a visual analysis made by four reviewers. Additionally, long-term image quality trends of a full-field digital mammography system and a computed radiography mammography system were investigated. Results The automated software produced feature detection levels comparable to visual analysis. The agreement was good in the case of fibers, while the software detected somewhat more microcalcifications and characteristic masses. Long-term follow-up via a quality assurance web portal demonstrated the feasibility of using the software for monitoring the performance of mammography systems in a multi-unit imaging center. Conclusion Automated image quality analysis enables monitoring the performance of digital mammography systems in an efficient, centralized manner.

  14. TH-B-207B-00: Pediatric Image Quality Optimization

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    NONE

    This imaging educational program will focus on solutions to common pediatric image quality optimization challenges. The speakers will present collective knowledge on best practices in pediatric imaging from their experience at dedicated children’s hospitals. One of the most commonly encountered pediatric imaging requirements for the non-specialist hospital is pediatric CT in the emergency room setting. Thus, this educational program will begin with optimization of pediatric CT in the emergency department. Though pediatric cardiovascular MRI may be less common in the non-specialist hospitals, low pediatric volumes and unique cardiovascular anatomy make optimization of these techniques difficult. Therefore, our second speaker willmore » review best practices in pediatric cardiovascular MRI based on experiences from a children’s hospital with a large volume of cardiac patients. Learning Objectives: To learn techniques for optimizing radiation dose and image quality for CT of children in the emergency room setting. To learn solutions for consistently high quality cardiovascular MRI of children.« less

  15. Image quality enhancement for skin cancer optical diagnostics

    NASA Astrophysics Data System (ADS)

    Bliznuks, Dmitrijs; Kuzmina, Ilona; Bolocko, Katrina; Lihachev, Alexey

    2017-12-01

    The research presents image quality analysis and enhancement proposals in biophotonic area. The sources of image problems are reviewed and analyzed. The problems with most impact in biophotonic area are analyzed in terms of specific biophotonic task - skin cancer diagnostics. The results point out that main problem for skin cancer analysis is the skin illumination problems. Since it is often not possible to prevent illumination problems, the paper proposes image post processing algorithm - low frequency filtering. Practical results show diagnostic results improvement after using proposed filter. Along that, filter do not reduces diagnostic results' quality for images without illumination defects. Current filtering algorithm requires empirical tuning of filter parameters. Further work needed to test the algorithm in other biophotonic applications and propose automatic filter parameter selection.

  16. Dependence of image quality on image operator and noise for optical diffusion tomography

    NASA Astrophysics Data System (ADS)

    Chang, Jenghwa; Graber, Harry L.; Barbour, Randall L.

    1998-04-01

    By applying linear perturbation theory to the radiation transport equation, the inverse problem of optical diffusion tomography can be reduced to a set of linear equations, W(mu) equals R, where W is the weight function, (mu) are the cross- section perturbations to be imaged, and R is the detector readings perturbations. We have studied the dependence of image quality on added systematic error and/or random noise in W and R. Tomographic data were collected from cylindrical phantoms, with and without added inclusions, using Monte Carlo methods. Image reconstruction was accomplished using a constrained conjugate gradient descent method. Result show that accurate images containing few artifacts are obtained when W is derived from a reference states whose optical thickness matches that of the unknown teste medium. Comparable image quality was also obtained for unmatched W, but the location of the target becomes more inaccurate as the mismatch increases. Results of the noise study show that image quality is much more sensitive to noise in W than in R, and the impact of noise increase with the number of iterations. Images reconstructed after pure noise was substituted for R consistently contain large peaks clustered about the cylinder axis, which was an initially unexpected structure. In other words, random input produces a non- random output. This finding suggests that algorithms sensitive to the evolution of this feature could be developed to suppress noise effects.

  17. Machine-Specific Magnetic Resonance Imaging Quality Control Procedures for Stereotactic Radiosurgery Treatment Planning

    PubMed Central

    Taghizadeh, Somayeh; Yang, Claus Chunli; R. Kanakamedala, Madhava; Morris, Bart; Vijayakumar, Srinivasan

    2017-01-01

    Purpose Magnetic resonance (MR) images are necessary for accurate contouring of intracranial targets, determination of gross target volume and evaluation of organs at risk during stereotactic radiosurgery (SRS) treatment planning procedures. Many centers use magnetic resonance imaging (MRI) simulators or regular diagnostic MRI machines for SRS treatment planning; while both types of machine require two stages of quality control (QC), both machine- and patient-specific, before use for SRS, no accepted guidelines for such QC currently exist. This article describes appropriate machine-specific QC procedures for SRS applications. Methods and materials We describe the adaptation of American College of Radiology (ACR)-recommended QC tests using an ACR MRI phantom for SRS treatment planning. In addition, commercial Quasar MRID3D and Quasar GRID3D phantoms were used to evaluate the effects of static magnetic field (B0) inhomogeneity, gradient nonlinearity, and a Leksell G frame (SRS frame) and its accessories on geometrical distortion in MR images. Results QC procedures found in-plane distortions (Maximum = 3.5 mm, Mean = 0.91 mm, Standard deviation = 0.67 mm, >2.5 mm (%) = 2) in X-direction (Maximum = 2.51 mm, Mean = 0.52 mm, Standard deviation = 0.39 mm, > 2.5 mm (%) = 0) and in Y-direction (Maximum = 13. 1 mm , Mean = 2.38 mm, Standard deviation = 2.45 mm, > 2.5 mm (%) = 34) in Z-direction and < 1 mm distortion at a head-sized region of interest. MR images acquired using a Leksell G frame and localization devices showed a mean absolute deviation of 2.3 mm from isocenter. The results of modified ACR tests were all within recommended limits, and baseline measurements have been defined for regular weekly QC tests. Conclusions With appropriate QC procedures in place, it is possible to routinely obtain clinically useful MR images suitable for SRS treatment planning purposes. MRI examination for SRS planning can benefit from the improved localization and planning possible

  18. Machine-Specific Magnetic Resonance Imaging Quality Control Procedures for Stereotactic Radiosurgery Treatment Planning.

    PubMed

    Fatemi, Ali; Taghizadeh, Somayeh; Yang, Claus Chunli; R Kanakamedala, Madhava; Morris, Bart; Vijayakumar, Srinivasan

    2017-12-18

    Purpose Magnetic resonance (MR) images are necessary for accurate contouring of intracranial targets, determination of gross target volume and evaluation of organs at risk during stereotactic radiosurgery (SRS) treatment planning procedures. Many centers use magnetic resonance imaging (MRI) simulators or regular diagnostic MRI machines for SRS treatment planning; while both types of machine require two stages of quality control (QC), both machine- and patient-specific, before use for SRS, no accepted guidelines for such QC currently exist. This article describes appropriate machine-specific QC procedures for SRS applications. Methods and materials We describe the adaptation of American College of Radiology (ACR)-recommended QC tests using an ACR MRI phantom for SRS treatment planning. In addition, commercial Quasar MRID 3D and Quasar GRID 3D phantoms were used to evaluate the effects of static magnetic field (B 0 ) inhomogeneity, gradient nonlinearity, and a Leksell G frame (SRS frame) and its accessories on geometrical distortion in MR images. Results QC procedures found in-plane distortions (Maximum = 3.5 mm, Mean = 0.91 mm, Standard deviation = 0.67 mm, >2.5 mm (%) = 2) in X-direction (Maximum = 2.51 mm, Mean = 0.52 mm, Standard deviation = 0.39 mm, > 2.5 mm (%) = 0) and in Y-direction (Maximum = 13. 1 mm , Mean = 2.38 mm, Standard deviation = 2.45 mm, > 2.5 mm (%) = 34) in Z-direction and < 1 mm distortion at a head-sized region of interest. MR images acquired using a Leksell G frame and localization devices showed a mean absolute deviation of 2.3 mm from isocenter. The results of modified ACR tests were all within recommended limits, and baseline measurements have been defined for regular weekly QC tests. Conclusions With appropriate QC procedures in place, it is possible to routinely obtain clinically useful MR images suitable for SRS treatment planning purposes. MRI examination for SRS planning can benefit from the improved localization and planning

  19. Body Image and Quality of Life in Adolescents With Craniofacial Conditions

    PubMed Central

    Crerand, Canice E.; Sarwer, David B.; Kazak, Anne E.; Clarke, Alexandra; DPsych; Rumsey, Nichola

    2017-01-01

    Objective To evaluate body image in adolescents with and without craniofacial conditions; and to examine relationships between body image and quality of life. Design Case-control design. Setting A pediatric hospital’s craniofacial center and primary care practices. Participants 70 adolescents with visible craniofacial conditions and a demographically-matched sample of 42 adolescents without craniofacial conditions. Main Outcome Measure Adolescents completed measures of quality of life and body image including satisfaction with weight, facial and overall appearance; investment in appearance (importance of appearance to self-worth); and body image disturbance (appearance-related distress and impairment in functioning). Results Adolescents with craniofacial conditions reported lower appearance investment (p < 0.001) and were more likely to report concerns about facial features (p < 0.02) compared to non-affected youth. Females in both groups reported greater investment in appearance, greater body image disturbance, and lower weight satisfaction compared to males (p < 0.01). Within both groups, greater body image disturbance was associated with lower quality of life (p <0.01). The two groups did not differ significantly on measures of quality of life, body image disturbance, or satisfaction with appearance. Conclusions Body image and quality of life in adolescents with craniofacial conditions are similar to non-affected youth. Relationships between body image and quality of life emphasize that appearance perceptions are important to adolescents’ well-being regardless of whether they have a facial disfigurement. Investment in one’s appearance may explain variations in body image satisfaction and serve as an intervention target particularly for females. PMID:26751907

  20. Image Quality Assessment Using the Joint Spatial/Spatial-Frequency Representation

    NASA Astrophysics Data System (ADS)

    Beghdadi, Azeddine; Iordache, Răzvan

    2006-12-01

    This paper demonstrates the usefulness of spatial/spatial-frequency representations in image quality assessment by introducing a new image dissimilarity measure based on 2D Wigner-Ville distribution (WVD). The properties of 2D WVD are shortly reviewed, and the important issue of choosing the analytic image is emphasized. The WVD-based measure is shown to be correlated with subjective human evaluation, which is the premise towards an image quality assessor developed on this principle.

  1. The influence of tinnitus acceptance on the quality of life and psychological distress in patients with chronic tinnitus

    PubMed Central

    Riedl, David; Rumpold, Gerhard; Schmidt, Annette; Zorowka, Patrick G.; Bliem, Harald R.; Moschen, Roland

    2015-01-01

    Recent findings show the importance of acceptance in the treatment of chronic tinnitus. So far, very limited research investigating the different levels of tinnitus acceptance has been conducted. The aim of this study was to investigate the quality of life (QoL) and psychological distress in patients with chronic tinnitus who reported different levels of tinnitus acceptance. The sample consisted of outpatients taking part in a tinnitus coping group (n = 97). Correlations between tinnitus acceptance, psychological distress, and QoL were calculated. Receiver operating characteristic (ROC) curves were used to calculate a cutoff score for the German “Tinnitus Acceptance Questionnaire” (CTAQ-G) and to evaluate the screening abilities of the CTAQ-G. Independent sample t-tests were conducted to compare QoL and psychological distress in patients with low tinnitus acceptance and high tinnitus acceptance. A cutoff point for CTAQ-G of 62.5 was defined, differentiating between patients with “low-to-mild tinnitus acceptance” and “moderate-to-high tinnitus acceptance.” Patients with higher levels of tinnitus acceptance reported a significantly higher QoL and lower psychological distress. Tinnitus acceptance plays an important role for patients with chronic tinnitus. Increased levels of acceptance are related to better QoL and less psychological distress. PMID:26356381

  2. Don't forget the posters! Quality and content variables associated with accepted abstracts at a national trauma meeting.

    PubMed

    Dossett, Lesly A; Fox, Erin E; del Junco, Deborah J; Zaydfudim, Victor; Kauffmann, Rondi; Shelton, Julia; Wang, Weiwei; Cioffi, William G; Holcomb, John B; Cotton, Bryan A

    2012-05-01

    As a primary venue for presenting research results, abstracts selected for presentation at national meetings should be of the highest scientific merit and research quality. It is uncertain to what degree this is achieved as the methodological quality of abstracts submitted to national surgical meetings has not been previously described. The objective of this study was to evaluate abstracts presented at a leading trauma meeting for methodological quality. All abstracts accepted for the 2009 American Association for the Surgery of Trauma meeting were reviewed and scored for methodological quality based on 10 criteria (scores, 0-10; 10 being the highest). Criteria were based on nationally published methodology guidelines. Two independent reviewers who were blinded to institution, region, and author reviewed each abstract. A total of 187 abstracts were accepted for presentation (67 oral and 120 posters). The most frequent clinical topics were shock/transfusion (23%), abdomen (12%), and nervous system (11%). Shock/transfusion abstracts were more common in the oral presentations (31% vs. 19%; p = 0.06). Abstracts from the northeast and south regions were the most common in both oral (26% and 29%) and posters (25% and 24%). Basic science accounted for 12% of accepted studies, while 51% were clinical and 28% were health services/outcomes. Only 8% of abstracts presented randomized data and only 11% reported null findings. Overall abstract scores ranged from 3 to 10 (median, 7; mean, 7.4). Abstracts selected for poster presentation had an overall higher score than those selected for oral presentation (7.4 ± 1.7 vs. 6.8 ± 1.7; p = 0.02). Although oral presentations traditionally receive the most attention and interest, the methodological quality of abstracts accepted for poster presentation equals (and sometimes exceeds) that of oral abstracts. Attendees of these national meetings should reconsider their time spent in viewing and visiting these poster sessions as with the

  3. Determination of pasture quality using airborne hyperspectral imaging

    NASA Astrophysics Data System (ADS)

    Pullanagari, R. R.; Kereszturi, G.; Yule, Ian J.; Irwin, M. E.

    2015-10-01

    Pasture quality is a critical determinant which influences animal performance (live weight gain, milk and meat production) and animal health. Assessment of pasture quality is therefore required to assist farmers with grazing planning and management, benchmarking between seasons and years. Traditionally, pasture quality is determined by field sampling which is laborious, expensive and time consuming, and the information is not available in real-time. Hyperspectral remote sensing has potential to accurately quantify biochemical composition of pasture over wide areas in great spatial detail. In this study an airborne imaging spectrometer (AisaFENIX, Specim) was used with a spectral range of 380-2500 nm with 448 spectral bands. A case study of a 600 ha hill country farm in New Zealand is used to illustrate the use of the system. Radiometric and atmospheric corrections, along with automatized georectification of the imagery using Digital Elevation Model (DEM), were applied to the raw images to convert into geocoded reflectance images. Then a multivariate statistical method, partial least squares (PLS), was applied to estimate pasture quality such as crude protein (CP) and metabolisable energy (ME) from canopy reflectance. The results from this study revealed that estimates of CP and ME had a R2 of 0.77 and 0.79, and RMSECV of 2.97 and 0.81 respectively. By utilizing these regression models, spatial maps were created over the imaged area. These pasture quality maps can be used for adopting precision agriculture practices which improves farm profitability and environmental sustainability.

  4. PROPELLER technique to improve image quality of MRI of the shoulder.

    PubMed

    Dietrich, Tobias J; Ulbrich, Erika J; Zanetti, Marco; Fucentese, Sandro F; Pfirrmann, Christian W A

    2011-12-01

    The purpose of this article is to evaluate the use of the periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) technique for artifact reduction and overall image quality improvement for intermediate-weighted and T2-weighted MRI of the shoulder. One hundred eleven patients undergoing MR arthrography of the shoulder were included. A coronal oblique intermediate-weighted turbo spin-echo (TSE) sequence with fat suppression and a sagittal oblique T2-weighted TSE sequence with fat suppression were obtained without (standard) and with the PROPELLER technique. Scanning time increased from 3 minutes 17 seconds to 4 minutes 17 seconds (coronal oblique plane) and from 2 minutes 52 seconds to 4 minutes 10 seconds (sagittal oblique) using PROPELLER. Two radiologists graded image artifacts, overall image quality, and delineation of several anatomic structures on a 5-point scale (5, no artifact, optimal diagnostic quality; and 1, severe artifacts, diagnostically not usable). The Wilcoxon signed rank test was used to compare the data of the standard and PROPELLER images. Motion artifacts were significantly reduced in PROPELLER images (p < 0.001). Observer 1 rated motion artifacts with diagnostic impairment in one patient on coronal oblique PROPELLER images compared with 33 patients on standard images. Ratings for the sequences with PROPELLER were significantly better for overall image quality (p < 0.001). Observer 1 noted an overall image quality with diagnostic impairment in nine patients on sagittal oblique PROPELLER images compared with 23 patients on standard MRI. The PROPELLER technique for MRI of the shoulder reduces the number of sequences with diagnostic impairment as a result of motion artifacts and increases image quality compared with standard TSE sequences. PROPELLER sequences increase the acquisition time.

  5. CAMEL: concept annotated image libraries

    NASA Astrophysics Data System (ADS)

    Natsev, Apostol; Chadha, Atul; Soetarman, Basuki; Vitter, Jeffrey S.

    2001-01-01

    The problem of content-based image searching has received considerable attention in the last few years. Thousands of images are now available on the Internet, and many important applications require searching of images in domains such as E-commerce, medical imaging, weather prediction, satellite imagery, and so on. Yet, content-based image querying is still largely unestablished as a mainstream field, nor is it widely used by search engines. We believe that two of the major hurdles for this poor acceptance are poor retrieval quality and usability.

  6. CAMEL: concept annotated image libraries

    NASA Astrophysics Data System (ADS)

    Natsev, Apostol; Chadha, Atul; Soetarman, Basuki; Vitter, Jeffrey S.

    2000-12-01

    The problem of content-based image searching has received considerable attention in the last few years. Thousands of images are now available on the Internet, and many important applications require searching of images in domains such as E-commerce, medical imaging, weather prediction, satellite imagery, and so on. Yet, content-based image querying is still largely unestablished as a mainstream field, nor is it widely used by search engines. We believe that two of the major hurdles for this poor acceptance are poor retrieval quality and usability.

  7. Prediction of Viking lander camera image quality

    NASA Technical Reports Server (NTRS)

    Huck, F. O.; Burcher, E. E.; Jobson, D. J.; Wall, S. D.

    1976-01-01

    Formulations are presented that permit prediction of image quality as a function of camera performance, surface radiance properties, and lighting and viewing geometry. Predictions made for a wide range of surface radiance properties reveal that image quality depends strongly on proper camera dynamic range command and on favorable lighting and viewing geometry. Proper camera dynamic range commands depend mostly on the surface albedo that will be encountered. Favorable lighting and viewing geometries depend mostly on lander orientation with respect to the diurnal sun path over the landing site, and tend to be independent of surface albedo and illumination scattering function. Side lighting with low sun elevation angles (10 to 30 deg) is generally favorable for imaging spatial details and slopes, whereas high sun elevation angles are favorable for measuring spectral reflectances.

  8. Synthesized view comparison method for no-reference 3D image quality assessment

    NASA Astrophysics Data System (ADS)

    Luo, Fangzhou; Lin, Chaoyi; Gu, Xiaodong; Ma, Xiaojun

    2018-04-01

    We develop a no-reference image quality assessment metric to evaluate the quality of synthesized view rendered from the Multi-view Video plus Depth (MVD) format. Our metric is named Synthesized View Comparison (SVC), which is designed for real-time quality monitoring at the receiver side in a 3D-TV system. The metric utilizes the virtual views in the middle which are warped from left and right views by Depth-image-based rendering algorithm (DIBR), and compares the difference between the virtual views rendered from different cameras by Structural SIMilarity (SSIM), a popular 2D full-reference image quality assessment metric. The experimental results indicate that our no-reference quality assessment metric for the synthesized images has competitive prediction performance compared with some classic full-reference image quality assessment metrics.

  9. Improved Self-Acceptance, Quality of Life, and Stress Level from Participation in a Worksite Yoga Foundations Program: A Pilot Study.

    PubMed

    Casey, Lori J; Van Rooy, Kimberly M; Sutherland, Stephanie J; Jenkins, Sarah M; Rosedahl, Jordan K; Wood, Nadia G; Ebbert, Jon O; Lopez-Jimenez, Francisco; Egginton, Jason S; Sim, Leslie A; Clark, Matthew M

    2018-03-29

    Yoga is increasing in popularity in the United States and across the globe. However, most yoga programs are provided outside the worksite; although many companies offer worksite wellness programs, at present there is limited documentation regarding the potential benefits of participating in a worksite yoga program. Therefore, the purpose of this project was to examine the potential effect of a worksite yoga program on self-acceptance, quality of life, and perceived stress. A prospective cohort pilot study that examined a structured worksite yoga program was designed and tailored to individuals new to yoga. The 8-week Yoga Foundations program was conducted at an academic medical center's worksite wellness center with 86 subjects. Outcome measures were the 36-item Self-Acceptance Scale; a six-item quality-of-life measure that assesses overall, social, mental, physical, emotional, and spiritual well-being; and the ten-item Perceived Stress Scale. Participants demonstrated significant improvement in their overall self-acceptance ( p < 0.001), quality of life ( p < 0.001), and perceived stress ( p ≤ 0.001) levels. They also highly rated the yoga instructors and the weekly format of the program. Participation in a Yoga Foundations program was associated with improvements in self-acceptance, quality of life and stress levels in worksite wellness center members. Future studies should use randomized designs and examine other wellness domains to learn more about the potential benefits of worksite yoga programs.

  10. Radiologists' confidence in detecting abnormalities on chest images and their subjective judgments of image quality

    NASA Astrophysics Data System (ADS)

    King, Jill L.; Gur, David; Rockette, Howard E.; Curtin, Hugh D.; Obuchowski, Nancy A.; Thaete, F. Leland; Britton, Cynthia A.; Metz, Charles E.

    1991-07-01

    The relationship between subjective judgments of image quality for the performance of specific detection tasks and radiologists' confidence level in arriving at correct diagnoses was investigated in two studies in which 12 readers, using a total of three different display environments, interpreted a series of 300 PA chest images. The modalities used were conventional films, laser-printed films, and high-resolution CRT display of digitized images. For the detection of interstitial disease, nodules, and pneumothoraces, there was no statistically significant correlation (Spearman rho) between subjective ratings of quality and radiologists' confidence in detecting these abnormalities. However, in each study, for all modalities and all readers but one, a small but statistically significant correlation was found between the radiologists' ability to correctly and confidently rule out interstitial disease and their subjective ratings of image quality.

  11. Application of statistical classification methods for predicting the acceptability of well-water quality

    NASA Astrophysics Data System (ADS)

    Cameron, Enrico; Pilla, Giorgio; Stella, Fabio A.

    2018-06-01

    The application of statistical classification methods is investigated—in comparison also to spatial interpolation methods—for predicting the acceptability of well-water quality in a situation where an effective quantitative model of the hydrogeological system under consideration cannot be developed. In the example area in northern Italy, in particular, the aquifer is locally affected by saline water and the concentration of chloride is the main indicator of both saltwater occurrence and groundwater quality. The goal is to predict if the chloride concentration in a water well will exceed the allowable concentration so that the water is unfit for the intended use. A statistical classification algorithm achieved the best predictive performances and the results of the study show that statistical classification methods provide further tools for dealing with groundwater quality problems concerning hydrogeological systems that are too difficult to describe analytically or to simulate effectively.

  12. Comparison of image quality and radiation exposure from C-arm fluoroscopes when used for imaging the spine.

    PubMed

    Prasarn, Mark L; Coyne, Ellen; Schreck, Michael; Rodgers, Jamie D; Rechtine, Glenn R

    2013-07-15

    Cadaveric imaging study. We sought to compare the fluoroscopic images produced by 4 different fluoroscopes for image quality and radiation exposure when used for imaging the spine. There are no previous published studies comparing mobile C-arm machines commonly used in clinical practice for imaging the spine. Anterior-posterior and lateral images of the cervical, thoracic, and lumbar spine were obtained from a cadaver placed supine on a radiolucent table. The fluoroscopy units used for the study included (1) GE OEC 9900 Elite (2010 model; General Electric Healthcare, Waukesha, WI), (2) Philips BV Pulsera (2009 model; Philips Healthcare, Andover, MA), (3) Philips BV Pulsera (2010 model; Philips Healthcare, Andover, MA), and (4) Siemens Arcadis Avantic (2010 model; Siemens Medical Solutions, Malvern, PA). The images were then downloaded, placed into a randomizer program, and evaluated by a group of spine surgeons and neuroradiologists independently. The reviewers, who were blinded to the fluoroscope the images were from, ranked them from best to worst using a numeric system. In addition, the images were rated according to a quality scale from 1 to 5, with 1 representing the best image quality. The radiation exposure level for the fluoroscopy units was also compared and was based on energy emission. According to the mean values for rank, the following order of best to worst was observed: (1) GE OEC > (2) Philips 2010 > (3) Philips 2009 > (4) Siemans. The exact same order was found when examining the image quality ratings. When comparing the radiation exposure level difference, it was observed that the OEC was the lowest, and there was a minimum 30% decrease in energy emission from the OEC versus the other C-arms studied. This is the first time that the spine image quality and radiation exposure of commonly used C-arm machines have been compared. The OEC was ranked the best, produced the best quality images, and had the least amount of radiation.

  13. Evaluation of image quality of MRI data for brain tumor surgery

    NASA Astrophysics Data System (ADS)

    Heckel, Frank; Arlt, Felix; Geisler, Benjamin; Zidowitz, Stephan; Neumuth, Thomas

    2016-03-01

    3D medical images are important components of modern medicine. Their usefulness for the physician depends on their quality, though. Only high-quality images allow accurate and reproducible diagnosis and appropriate support during treatment. We have analyzed 202 MRI images for brain tumor surgery in a retrospective study. Both an experienced neurosurgeon and an experienced neuroradiologist rated each available image with respect to its role in the clinical workflow, its suitability for this specific role, various image quality characteristics, and imaging artifacts. Our results show that MRI data acquired for brain tumor surgery does not always fulfill the required quality standards and that there is a significant disagreement between the surgeon and the radiologist, with the surgeon being more critical. Noise, resolution, as well as the coverage of anatomical structures were the most important criteria for the surgeon, while the radiologist was mainly disturbed by motion artifacts.

  14. Improving image quality in laboratory x-ray phase-contrast imaging

    NASA Astrophysics Data System (ADS)

    De Marco, F.; Marschner, M.; Birnbacher, L.; Viermetz, M.; Noël, P.; Herzen, J.; Pfeiffer, F.

    2017-03-01

    Grating-based X-ray phase-contrast (gbPC) is known to provide significant benefits for biomedical imaging. To investigate these benefits, a high-sensitivity gbPC micro-CT setup for small (≍ 5 cm) biological samples has been constructed. Unfortunately, high differential-phase sensitivity leads to an increased magnitude of data processing artifacts, limiting the quality of tomographic reconstructions. Most importantly, processing of phase-stepping data with incorrect stepping positions can introduce artifacts resembling Moiré fringes to the projections. Additionally, the focal spot size of the X-ray source limits resolution of tomograms. Here we present a set of algorithms to minimize artifacts, increase resolution and improve visual impression of projections and tomograms from the examined setup. We assessed two algorithms for artifact reduction: Firstly, a correction algorithm exploiting correlations of the artifacts and differential-phase data was developed and tested. Artifacts were reliably removed without compromising image data. Secondly, we implemented a new algorithm for flatfield selection, which was shown to exclude flat-fields with strong artifacts. Both procedures successfully improved image quality of projections and tomograms. Deconvolution of all projections of a CT scan can minimize blurring introduced by the finite size of the X-ray source focal spot. Application of the Richardson-Lucy deconvolution algorithm to gbPC-CT projections resulted in an improved resolution of phase-contrast tomograms. Additionally, we found that nearest-neighbor interpolation of projections can improve the visual impression of very small features in phase-contrast tomograms. In conclusion, we achieved an increase in image resolution and quality for the investigated setup, which may lead to an improved detection of very small sample features, thereby maximizing the setup's utility.

  15. Image quality specification and maintenance for airborne SAR

    NASA Astrophysics Data System (ADS)

    Clinard, Mark S.

    2004-08-01

    Specification, verification, and maintenance of image quality over the lifecycle of an operational airborne SAR begin with the specification for the system itself. Verification of image quality-oriented specification compliance can be enhanced by including a specification requirement that a vendor provide appropriate imagery at the various phases of the system life cycle. The nature and content of the imagery appropriate for each stage of the process depends on the nature of the test, the economics of collection, and the availability of techniques to extract the desired information from the data. At the earliest lifecycle stages, Concept and Technology Development (CTD) and System Development and Demonstration (SDD), the test set could include simulated imagery to demonstrate the mathematical and engineering concepts being implemented thus allowing demonstration of compliance, in part, through simulation. For Initial Operational Test and Evaluation (IOT&E), imagery collected from precisely instrumented test ranges and targets of opportunity consisting of a priori or a posteriori ground-truthed cultural and natural features are of value to the analysis of product quality compliance. Regular monitoring of image quality is possible using operational imagery and automated metrics; more precise measurements can be performed with imagery of instrumented scenes, when available. A survey of image quality measurement techniques is presented along with a discussion of the challenges of managing an airborne SAR program with the scarce resources of time, money, and ground-truthed data. Recommendations are provided that should allow an improvement in the product quality specification and maintenance process with a minimal increase in resource demands on the customer, the vendor, the operational personnel, and the asset itself.

  16. Maternal diet during early childhood, but not pregnancy, predicts diet quality and fruit and vegetable acceptance in offspring.

    PubMed

    Ashman, Amy M; Collins, Clare E; Hure, Alexis J; Jensen, Megan; Oldmeadow, Christopher

    2016-07-01

    Studies have identified prenatal flavour exposure as a determinant of taste preferences in infants; however, these studies have focused on relatively small samples and limited flavours. As many parents struggle with getting children to accept a variety of nutritious foods, a study of the factors influencing food acceptance is warranted. The objective of this study was to determine whether exposure to a wider variety of fruit and vegetables and overall higher diet quality in utero results in acceptance of a greater variety of these foods and better diet quality for offspring during childhood. This study is a secondary data analysis of pregnant women (n = 52) and their resulting offspring recruited for the Women and Their Children's Health study in NSW, Australia. Dietary intake of mothers and children was measured using food frequency questionnaires. Diet quality and vegetable and fruit variety were calculated using the Australian Recommended Food Score and the Australian Child and Adolescent Recommended Food Score. Associations between maternal and child diet quality and variety were assessed using Pearson's correlations and the total effect of in utero maternal pregnancy diet on childhood diet was decomposed into direct and indirect effect using mediation analysis. Maternal pregnancy and post-natal diet were both correlated with child diet for overall diet quality and fruit and vegetable variety (P < 0.001). Mediation analyses showed that the indirect effect of maternal pregnancy diet on child diet was mediated through maternal post-natal diet, particularly for fruit (P = 0.045) and vegetables (P = 0.055). Nutrition intervention should therefore be aimed at improving diet quality and variety in mothers with young children, in order to subsequently improve eating habits of offspring. © 2014 John Wiley & Sons Ltd.

  17. Acceptability of the Conceptions of Higher Education Quality to First Year Students of the Study Field of Pedagogy

    ERIC Educational Resources Information Center

    Žibeniene, Gintaute; Savickiene, Izabela

    2014-01-01

    The article presents which conceptions of higher education quality are most acceptable to first-year students of the study field of pedagogy. It is significant to analyse students' opinions as more than 10 years ago the EU member states agreed that higher education institutions bear responsibility for the quality of higher education. Being members…

  18. Study of a water quality imager for coastal zone missions

    NASA Technical Reports Server (NTRS)

    Staylor, W. F.; Harrison, E. F.; Wessel, V. W.

    1975-01-01

    The present work surveys water quality user requirements and then determines the general characteristics of an orbiting imager (the Applications Explorer, or AE) dedicated to the measurement of water quality, which could be used as a low-cost means of testing advanced imager concepts and assessing the ability of imager techniques to meet the goals of a comprehensive water quality monitoring program. The proposed imager has four spectral bands, a spatial resolution of 25 meters, and swath width of 36 km with a pointing capability of 330 km. Silicon photodetector arrays, pointing systems, and several optical features are included. A nominal orbit of 500 km altitude at an inclination of 50 deg is recommended.

  19. Maternal relationship quality and peer social acceptance as mediators between child maltreatment and adolescent depressive symptoms: Gender differences.

    PubMed

    Alto, Michelle; Handley, Elizabeth; Rogosch, Fred; Cicchetti, Dante; Toth, Sheree

    2018-02-01

    Childhood maltreatment negatively impacts the development of maternal and peer relationships and may put adolescents at risk for depression. The present study examined gender differences in maternal relationship quality and peer social acceptance as mediators of the association between childhood maltreatment and adolescent depressive symptoms in 342 (151 female, 191 male) maltreated (n = 198) and nonmaltreated (n = 144) youth in the USA. An observer report Q-Scale measure of depressive symptoms was developed and received preliminary support. Social acceptance was a significant mediator for both genders. The significant association between maltreatment and maternal relationship quality was unique to females, and the association between maternal relationship quality and depressive symptoms was significantly stronger for females. Lower maternal relationship quality marginally significantly mediated the association between maltreatment and depressive symptoms for females only. Results have implications for the prevention and intervention of depression in adolescents with a history of childhood maltreatment. Copyright © 2017 The Foundation for Professionals in Services for Adolescents. Published by Elsevier Ltd. All rights reserved.

  20. High-field open versus short-bore magnetic resonance imaging of the spine: a randomized controlled comparison of image quality.

    PubMed

    Enders, Judith; Rief, Matthias; Zimmermann, Elke; Asbach, Patrick; Diederichs, Gerd; Wetz, Christoph; Siebert, Eberhard; Wagner, Moritz; Hamm, Bernd; Dewey, Marc

    2013-01-01

    The purpose of the present study was to compare the image quality of spinal magnetic resonance (MR) imaging performed on a high-field horizontal open versus a short-bore MR scanner in a randomized controlled study setup. Altogether, 93 (80% women, mean age 53) consecutive patients underwent spine imaging after random assignement to a 1-T horizontal open MR scanner with a vertical magnetic field or a 1.5-T short-bore MR scanner. This patient subset was part of a larger cohort. Image quality was assessed by determining qualitative parameters, signal-to-noise (SNR) and contrast-to-noise ratios (CNR), and quantitative contour sharpness. The image quality parameters were higher for short-bore MR imaging. Regarding all sequences, the relative differences were 39% for the mean overall qualitative image quality, 53% for the mean SNR values, and 34-37% for the quantitative contour sharpness (P<0.0001). The CNR values were also higher for images obtained with the short-bore MR scanner. No sequence was of very poor (nondiagnostic) image quality. Scanning times were significantly longer for examinations performed on the open MR scanner (mean: 32±22 min versus 20±9 min; P<0.0001). In this randomized controlled comparison of spinal MR imaging with an open versus a short-bore scanner, short-bore MR imaging revealed considerably higher image quality with shorter scanning times. ClinicalTrials.gov NCT00715806.

  1. High-Field Open versus Short-Bore Magnetic Resonance Imaging of the Spine: A Randomized Controlled Comparison of Image Quality

    PubMed Central

    Zimmermann, Elke; Asbach, Patrick; Diederichs, Gerd; Wetz, Christoph; Siebert, Eberhard; Wagner, Moritz; Hamm, Bernd; Dewey, Marc

    2013-01-01

    Background The purpose of the present study was to compare the image quality of spinal magnetic resonance (MR) imaging performed on a high-field horizontal open versus a short-bore MR scanner in a randomized controlled study setup. Methods Altogether, 93 (80% women, mean age 53) consecutive patients underwent spine imaging after random assignement to a 1-T horizontal open MR scanner with a vertical magnetic field or a 1.5-T short-bore MR scanner. This patient subset was part of a larger cohort. Image quality was assessed by determining qualitative parameters, signal-to-noise (SNR) and contrast-to-noise ratios (CNR), and quantitative contour sharpness. Results The image quality parameters were higher for short-bore MR imaging. Regarding all sequences, the relative differences were 39% for the mean overall qualitative image quality, 53% for the mean SNR values, and 34–37% for the quantitative contour sharpness (P<0.0001). The CNR values were also higher for images obtained with the short-bore MR scanner. No sequence was of very poor (nondiagnostic) image quality. Scanning times were significantly longer for examinations performed on the open MR scanner (mean: 32±22 min versus 20±9 min; P<0.0001). Conclusions In this randomized controlled comparison of spinal MR imaging with an open versus a short-bore scanner, short-bore MR imaging revealed considerably higher image quality with shorter scanning times. Trial Registration ClinicalTrials.gov NCT00715806 PMID:24391767

  2. Body image and quality of life in a Spanish population

    PubMed Central

    Lobera, Ignacio Jáuregui; Ríos, Patricia Bolaños

    2011-01-01

    Purpose The aim of the current study was to analyze the psychometric properties, factor structure, and internal consistency of the Spanish version of the Body Image Quality of Life Inventory (BIQLI-SP) as well as its test–retest reliability. Further objectives were to analyze different relationships with key dimensions of psychosocial functioning (ie, self-esteem, presence of psychopathological symptoms, eating and body image-related problems, and perceived stress) and to evaluate differences in body image quality of life due to gender. Patients and methods The sample comprised 417 students without any psychiatric history, recruited from the Pablo de Olavide University and the University of Seville. There were 140 men (33.57%) and 277 women (66.43%), and the mean age was 21.62 years (standard deviation = 5.12). After obtaining informed consent from all participants, the following questionnaires were administered: BIQLI, Eating Disorder Inventory-2 (EDI-2), Perceived Stress Questionnaire (PSQ), Self-Esteem Scale (SES), and Symptom Checklist-90-Revised (SCL-90-R). Results The BIQLI-SP shows adequate psychometric properties, and it may be useful to determine the body image quality of life in different physical conditions. A more positive body image quality of life is associated with better self-esteem, better psychological wellbeing, and fewer eating-related dysfunctional attitudes, this being more evident among women. Conclusion The BIQLI-SP may be useful to determine the body image quality of life in different contexts with regard to dermatology, cosmetic and reconstructive surgery, and endocrinology, among others. In these fields of study, a new trend has emerged to assess body image-related quality of life. PMID:21403794

  3. High-resolution T2-weighted abdominal magnetic resonance imaging using respiratory triggering: impact of butylscopolamine on image quality.

    PubMed

    Wagner, M; Klessen, C; Rief, M; Elgeti, T; Taupitz, M; Hamm, B; Asbach, P

    2008-05-01

    Respiratory triggering allows the acquisition of high-resolution magnetic resonance (MR) images of the upper abdomen. However, the depiction of organs close to the gastrointestinal tract can be considerably impaired by ghosting artifacts and blurring caused by bowel peristalsis. To evaluate the effect of gastrointestinal motion suppression by intramuscular butylscopolamine administration on the image quality of a respiratory-triggered T2-weighted turbo spin-echo (T2w TSE) sequence of the upper abdomen. Images of 46 patients were retrospectively analyzed. Twenty-four patients had received intramuscular injection of 40 mg butylscopolamine immediately before MR imaging. Fourteen of the 24 patients in the butylscopolamine group underwent repeat imaging after a mean of 29 min. Quantitative analysis of the ghosting artifacts was done by measuring signal intensities in regions of interest placed in air anterior to the patient. In addition, image quality was assessed qualitatively by two radiologists by consensus. Spasmolytic medication with butylscopolamine reduced ghosting artifacts and significantly improved image quality of the respiratory-triggered T2w TSE sequence. The most pronounced effect of butylscopolamine administration on image quality was found for the pancreas and the left hepatic lobe. The rate of examinations with excellent or good depiction of the pancreas and the left hepatic lobe in the group without premedication and in the butylscopolamine group was 55% vs. 96% (pancreatic head), 35% vs. 88% (pancreatic body), 43% vs. 96% (pancreatic tail), and 45% vs. 83% (left hepatic lobe), respectively. Regarding the duration of the effect of intramuscular butylscopolamine, repeat imaging after a mean of 29 min did not result in a significant deterioration of image quality. Intramuscular butylscopolamine administration significantly improves image quality of respiratory-triggered T2-weighted abdominal MR imaging by persistent reduction of peristaltic artifacts. MR

  4. Face detection on distorted images using perceptual quality-aware features

    NASA Astrophysics Data System (ADS)

    Gunasekar, Suriya; Ghosh, Joydeep; Bovik, Alan C.

    2014-02-01

    We quantify the degradation in performance of a popular and effective face detector when human-perceived image quality is degraded by distortions due to additive white gaussian noise, gaussian blur or JPEG compression. It is observed that, within a certain range of perceived image quality, a modest increase in image quality can drastically improve face detection performance. These results can be used to guide resource or bandwidth allocation in a communication/delivery system that is associated with face detection tasks. A new face detector based on QualHOG features is also proposed that augments face-indicative HOG features with perceptual quality-aware spatial Natural Scene Statistics (NSS) features, yielding improved tolerance against image distortions. The new detector provides statistically significant improvements over a strong baseline on a large database of face images representing a wide range of distortions. To facilitate this study, we created a new Distorted Face Database, containing face and non-face patches from images impaired by a variety of common distortion types and levels. This new dataset is available for download and further experimentation at www.ideal.ece.utexas.edu/˜suriya/DFD/.

  5. Optimizing radiation exposure in screening of body packing: image quality and diagnostic acceptability of an 80 kVp protocol with automated tube current modulation.

    PubMed

    Aissa, Joel; Boos, Johannes; Rubbert, Christian; Caspers, Julian; Schleich, Christoph; Thomas, Christoph; Kröpil, Patric; Antoch, Gerald; Miese, Falk

    2017-06-01

    The aim of this study was to evaluate the objective and subjective image quality of a novel computed tomography (CT) protocol with reduced radiation dose for body packing with 80 kVp and automated tube current modulation (ATCM) compared to a standard body packing CT protocol. 80 individuals who were examined between March 2012 and July 2015 in suspicion of ingested drug packets were retrospectively included in this study. Thirty-one CT examinations were performed using ATCM and a fixed tube voltage of 80 kVp (group A). Forty-nine CT examinations were performed using a standard protocol with a tube voltage of 120 kVp and a fixed tube current time product of 40 mAs (group B). Subjective and objective image quality and visibility of drug packets were assessed. Radiation exposure of both protocols was compared. Contrast-to-noise ratio (group A: 0.56 ± 0.36; group B: 1.13 ± 0.91) and Signal-to-noise ratio (group A: 3.69 ± 0.98; group B: 7.08 ± 2.67) were significantly lower for group A compared to group B (p < 0.001). Subjectively, image quality was decreased for group A compared to group B (2.5 ± 0.8 vs. 1.2 ± 0.4; p < 0.001). Attenuation of body packets was higher with the new protocol (group A: 362.2 ± 70.3 Hounsfield Units (HU); group B: 210.6 ± 60.2 HU; p = 0.005). Volumetric Computed Tomography Dose Index (CTDIvol) and Dose Length Product (DLP) were significantly lower in group A (CTDIvol 2.2 ± 0.9 mGy, DLP 105.7 ± 52.3 mGycm) as compared to group B (CTDIvol 2.7 ± 0.1 mGy, DLP 126.0 ± 9.7 mGycm, p = 0.002 and p = 0.01). The novel 80 kVp CT protocol with ATCM leads to a significant dose reduction compared to a standard CT body packing protocol. The novel protocol led to a diagnostic image quality and cocaine body packets were reliably detected due to the high attenuation.

  6. Radiation dose and image quality for paediatric interventional cardiology

    NASA Astrophysics Data System (ADS)

    Vano, E.; Ubeda, C.; Leyton, F.; Miranda, P.

    2008-08-01

    Radiation dose and image quality for paediatric protocols in a biplane x-ray system used for interventional cardiology have been evaluated. Entrance surface air kerma (ESAK) and image quality using a test object and polymethyl methacrylate (PMMA) phantoms have been measured for the typical paediatric patient thicknesses (4-20 cm of PMMA). Images from fluoroscopy (low, medium and high) and cine modes have been archived in digital imaging and communications in medicine (DICOM) format. Signal-to-noise ratio (SNR), figure of merit (FOM), contrast (CO), contrast-to-noise ratio (CNR) and high contrast spatial resolution (HCSR) have been computed from the images. Data on dose transferred to the DICOM header have been used to test the values of the dosimetric display at the interventional reference point. ESAK for fluoroscopy modes ranges from 0.15 to 36.60 µGy/frame when moving from 4 to 20 cm PMMA. For cine, these values range from 2.80 to 161.10 µGy/frame. SNR, FOM, CO, CNR and HCSR are improved for high fluoroscopy and cine modes and maintained roughly constant for the different thicknesses. Cumulative dose at the interventional reference point resulted 25-45% higher than the skin dose for the vertical C-arm (depending of the phantom thickness). ESAK and numerical image quality parameters allow the verification of the proper setting of the x-ray system. Knowing the increases in dose per frame when increasing phantom thicknesses together with the image quality parameters will help cardiologists in the good management of patient dose and allow them to select the best imaging acquisition mode during clinical procedures.

  7. Image Quality Improvement in Adaptive Optics Scanning Laser Ophthalmoscopy Assisted Capillary Visualization Using B-spline-based Elastic Image Registration

    PubMed Central

    Uji, Akihito; Ooto, Sotaro; Hangai, Masanori; Arichika, Shigeta; Yoshimura, Nagahisa

    2013-01-01

    Purpose To investigate the effect of B-spline-based elastic image registration on adaptive optics scanning laser ophthalmoscopy (AO-SLO)-assisted capillary visualization. Methods AO-SLO videos were acquired from parafoveal areas in the eyes of healthy subjects and patients with various diseases. After nonlinear image registration, the image quality of capillary images constructed from AO-SLO videos using motion contrast enhancement was compared before and after B-spline-based elastic (nonlinear) image registration performed using ImageJ. For objective comparison of image quality, contrast-to-noise ratios (CNRS) for vessel images were calculated. For subjective comparison, experienced ophthalmologists ranked images on a 5-point scale. Results All AO-SLO videos were successfully stabilized by elastic image registration. CNR was significantly higher in capillary images stabilized by elastic image registration than in those stabilized without registration. The average ratio of CNR in images with elastic image registration to CNR in images without elastic image registration was 2.10 ± 1.73, with no significant difference in the ratio between patients and healthy subjects. Improvement of image quality was also supported by expert comparison. Conclusions Use of B-spline-based elastic image registration in AO-SLO-assisted capillary visualization was effective for enhancing image quality both objectively and subjectively. PMID:24265796

  8. Optimized Plane Wave Imaging for Fast and High-Quality Ultrasound Imaging.

    PubMed

    Jensen, Jonas; Stuart, Matthias Bo; Jensen, Jorgen Arendt

    2016-11-01

    This paper presents a method for optimizing parameters affecting the image quality in plane wave imaging. More specifically, the number of emissions and steering angles is optimized to attain the best images with the highest frame rate possible. The method is applied to a specific problem, where image quality for a λ -pitch transducer is compared with a λ /2-pitch transducer. Grating lobe artifacts for λ -pitch transducers degrade the contrast in plane wave images, and the impact on frame rate is studied. Field II simulations of plane wave images are made for all combinations of the parameters, and the optimal setup is selected based on Pareto optimality. The optimal setup for a simulated 4.1-MHz λ -pitch transducer uses 61 emissions and a maximum steering angle of 20° for depths from 0 to 60 mm. The achieved lateral full-width at half-maximum (FWHM) is 1.5λ and the contrast is -29 dB for a scatterer at 9 mm ( 24λ ). Using a λ /2-pitch transducer and only 21 emissions within the same angle range, the image quality is improved in terms of contrast, which is -37 dB. For imaging in regions deeper than 25 mm ( 66λ ), only 21 emissions are optimal for both the transducers, resulting in a -36 dB contrast at 34 mm ( 90λ ). Measurements are performed using the experimental SARUS scanner connected to a λ -pitch and λ /2-pitch transducer. A wire phantom and a tissue mimicking phantom containing anechoic cysts are scanned and show the performance using the optimized sequences for the transducers. FWHM is 1.6λ and contrast is -25 dB for a wire at 9 mm using the λ -pitch transducer. For the λ /2-pitch transducer, contrast is -29 dB. In vivo scans of the carotid artery of a healthy volunteer show improved contrast and present fewer artifacts, when using the λ /2-pitch transducer compared with the λ -pitch. It is demonstrated with a frame rate, which is three times higher for the λ /2-pitch transducer.

  9. Defining Quality in Cardiovascular Imaging: A Scientific Statement From the American Heart Association.

    PubMed

    Shaw, Leslee J; Blankstein, Ron; Jacobs, Jill E; Leipsic, Jonathon A; Kwong, Raymond Y; Taqueti, Viviany R; Beanlands, Rob S B; Mieres, Jennifer H; Flamm, Scott D; Gerber, Thomas C; Spertus, John; Di Carli, Marcelo F

    2017-12-01

    The aims of the current statement are to refine the definition of quality in cardiovascular imaging and to propose novel methodological approaches to inform the demonstration of quality in imaging in future clinical trials and registries. We propose defining quality in cardiovascular imaging using an analytical framework put forth by the Institute of Medicine whereby quality was defined as testing being safe, effective, patient-centered, timely, equitable, and efficient. The implications of each of these components of quality health care are as essential for cardiovascular imaging as they are for other areas within health care. Our proposed statement may serve as the foundation for integrating these quality indicators into establishing designations of quality laboratory practices and developing standards for value-based payment reform for imaging services. We also include recommendations for future clinical research to fulfill quality aims within cardiovascular imaging, including clinical hypotheses of improving patient outcomes, the importance of health status as an end point, and deferred testing options. Future research should evolve to define novel methods optimized for the role of cardiovascular imaging for detecting disease and guiding treatment and to demonstrate the role of cardiovascular imaging in facilitating healthcare quality. © 2017 American Heart Association, Inc.

  10. Perceptual quality prediction on authentically distorted images using a bag of features approach

    PubMed Central

    Ghadiyaram, Deepti; Bovik, Alan C.

    2017-01-01

    Current top-performing blind perceptual image quality prediction models are generally trained on legacy databases of human quality opinion scores on synthetically distorted images. Therefore, they learn image features that effectively predict human visual quality judgments of inauthentic and usually isolated (single) distortions. However, real-world images usually contain complex composite mixtures of multiple distortions. We study the perceptually relevant natural scene statistics of such authentically distorted images in different color spaces and transform domains. We propose a “bag of feature maps” approach that avoids assumptions about the type of distortion(s) contained in an image and instead focuses on capturing consistencies—or departures therefrom—of the statistics of real-world images. Using a large database of authentically distorted images, human opinions of them, and bags of features computed on them, we train a regressor to conduct image quality prediction. We demonstrate the competence of the features toward improving automatic perceptual quality prediction by testing a learned algorithm using them on a benchmark legacy database as well as on a newly introduced distortion-realistic resource called the LIVE In the Wild Image Quality Challenge Database. We extensively evaluate the perceptual quality prediction model and algorithm and show that it is able to achieve good-quality prediction power that is better than other leading models. PMID:28129417

  11. Radiation dose optimization in pediatric temporal bone computed tomography: influence of tube tension on image contrast and image quality.

    PubMed

    Nauer, Claude Bertrand; Zubler, Christoph; Weisstanner, Christian; Stieger, Christof; Senn, Pascal; Arnold, Andreas

    2012-03-01

    The purpose of this experimental study was to investigate the effect of tube tension reduction on image contrast and image quality in pediatric temporal bone computed tomography (CT). Seven lamb heads with infant-equivalent sizes were scanned repeatedly, using four tube tensions from 140 to 80 kV while the CT-Dose Index (CTDI) was held constant. Scanning was repeated with four CTDI values from 30 to 3 mGy. Image contrast was calculated for the middle ear as the Hounsfield unit (HU) difference between bone and air and for the inner ear as the HU difference between bone and fluid. The influence of tube tension on high-contrast detail delineation was evaluated using a phantom. The subjective image quality of eight middle and inner ear structures was assessed using a 4-point scale (scores 1-2 = insufficient; scores 3-4 = sufficient). Middle and inner ear contrast showed a near linear increase with tube tension reduction (r = -0.94/-0.88) and was highest at 80 kV. Tube tension had no influence on spatial resolution. Subjective image quality analysis showed significantly better scoring at lower tube tensions, with highest image quality at 80 kV. However, image quality improvement was most relevant for low-dose scans. Image contrast in the temporal bone is significantly higher at low tube tensions, leading to a better subjective image quality. Highest contrast and best quality were found at 80 kV. This image quality improvement might be utilized to further reduce the radiation dose in pediatric low-dose CT protocols.

  12. Relationships between body image, sexual satisfaction, and relationship quality in romantic couples.

    PubMed

    van den Brink, Femke; Vollmann, Manja; Smeets, Monique A M; Hessen, David J; Woertman, Liesbeth

    2018-03-08

    Previous studies found important associations between body image, sexual satisfaction, and perceived romantic relationship quality, but mainly focused on one individual's perceptions rather than both partners. To take the interdependency of romantic partners into account, the present study examined these associations in romantic couples with a dyadic approach. In a cross-sectional design, 151 Dutch heterosexual couples completed an online survey measuring body image, sexual satisfaction, and perceived relationship quality. Hypotheses were tested using the actor-partner interdependence model (APIM) and an APIM extended with a mediator (APIMeM), with couple members' body image as predictors, couple members' sexual satisfaction as mediators, and couple members' relationship quality as outcomes. Results indicated that within individuals, a more positive body image was linked to higher perceived romantic relationship quality through greater sexual satisfaction. No gender differences were found, implying that body image and sexual satisfaction are equally strongly associated with perceived relationship quality in women and men. Results revealed no associations of an individual's body image and sexual satisfaction with the partner's perceived relationship quality. These findings implicate that interventions focusing on developing and maintaining a positive body image may be helpful in building on a more satisfying sex life and higher perceived relationship quality. (PsycINFO Database Record (c) 2018 APA, all rights reserved).

  13. Image quality of mixed convolution kernel in thoracic computed tomography.

    PubMed

    Neubauer, Jakob; Spira, Eva Maria; Strube, Juliane; Langer, Mathias; Voss, Christian; Kotter, Elmar

    2016-11-01

    The mixed convolution kernel alters his properties geographically according to the depicted organ structure, especially for the lung. Therefore, we compared the image quality of the mixed convolution kernel to standard soft and hard kernel reconstructions for different organ structures in thoracic computed tomography (CT) images.Our Ethics Committee approved this prospective study. In total, 31 patients who underwent contrast-enhanced thoracic CT studies were included after informed consent. Axial reconstructions were performed with hard, soft, and mixed convolution kernel. Three independent and blinded observers rated the image quality according to the European Guidelines for Quality Criteria of Thoracic CT for 13 organ structures. The observers rated the depiction of the structures in all reconstructions on a 5-point Likert scale. Statistical analysis was performed with the Friedman Test and post hoc analysis with the Wilcoxon rank-sum test.Compared to the soft convolution kernel, the mixed convolution kernel was rated with a higher image quality for lung parenchyma, segmental bronchi, and the border between the pleura and the thoracic wall (P < 0.03). Compared to the hard convolution kernel, the mixed convolution kernel was rated with a higher image quality for aorta, anterior mediastinal structures, paratracheal soft tissue, hilar lymph nodes, esophagus, pleuromediastinal border, large and medium sized pulmonary vessels and abdomen (P < 0.004) but a lower image quality for trachea, segmental bronchi, lung parenchyma, and skeleton (P < 0.001).The mixed convolution kernel cannot fully substitute the standard CT reconstructions. Hard and soft convolution kernel reconstructions still seem to be mandatory for thoracic CT.

  14. Quantitative metrics for assessment of chemical image quality and spatial resolution

    DOE PAGES

    Kertesz, Vilmos; Cahill, John F.; Van Berkel, Gary J.

    2016-02-28

    Rationale: Currently objective/quantitative descriptions of the quality and spatial resolution of mass spectrometry derived chemical images are not standardized. Development of these standardized metrics is required to objectively describe chemical imaging capabilities of existing and/or new mass spectrometry imaging technologies. Such metrics would allow unbiased judgment of intra-laboratory advancement and/or inter-laboratory comparison for these technologies if used together with standardized surfaces. Methods: We developed two image metrics, viz., chemical image contrast (ChemIC) based on signal-to-noise related statistical measures on chemical image pixels and corrected resolving power factor (cRPF) constructed from statistical analysis of mass-to-charge chronograms across features of interest inmore » an image. These metrics, quantifying chemical image quality and spatial resolution, respectively, were used to evaluate chemical images of a model photoresist patterned surface collected using a laser ablation/liquid vortex capture mass spectrometry imaging system under different instrument operational parameters. Results: The calculated ChemIC and cRPF metrics determined in an unbiased fashion the relative ranking of chemical image quality obtained with the laser ablation/liquid vortex capture mass spectrometry imaging system. These rankings were used to show that both chemical image contrast and spatial resolution deteriorated with increasing surface scan speed, increased lane spacing and decreasing size of surface features. Conclusions: ChemIC and cRPF, respectively, were developed and successfully applied for the objective description of chemical image quality and spatial resolution of chemical images collected from model surfaces using a laser ablation/liquid vortex capture mass spectrometry imaging system.« less

  15. Quantitative metrics for assessment of chemical image quality and spatial resolution

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kertesz, Vilmos; Cahill, John F.; Van Berkel, Gary J.

    Rationale: Currently objective/quantitative descriptions of the quality and spatial resolution of mass spectrometry derived chemical images are not standardized. Development of these standardized metrics is required to objectively describe chemical imaging capabilities of existing and/or new mass spectrometry imaging technologies. Such metrics would allow unbiased judgment of intra-laboratory advancement and/or inter-laboratory comparison for these technologies if used together with standardized surfaces. Methods: We developed two image metrics, viz., chemical image contrast (ChemIC) based on signal-to-noise related statistical measures on chemical image pixels and corrected resolving power factor (cRPF) constructed from statistical analysis of mass-to-charge chronograms across features of interest inmore » an image. These metrics, quantifying chemical image quality and spatial resolution, respectively, were used to evaluate chemical images of a model photoresist patterned surface collected using a laser ablation/liquid vortex capture mass spectrometry imaging system under different instrument operational parameters. Results: The calculated ChemIC and cRPF metrics determined in an unbiased fashion the relative ranking of chemical image quality obtained with the laser ablation/liquid vortex capture mass spectrometry imaging system. These rankings were used to show that both chemical image contrast and spatial resolution deteriorated with increasing surface scan speed, increased lane spacing and decreasing size of surface features. Conclusions: ChemIC and cRPF, respectively, were developed and successfully applied for the objective description of chemical image quality and spatial resolution of chemical images collected from model surfaces using a laser ablation/liquid vortex capture mass spectrometry imaging system.« less

  16. Quality and Acceptability of Meat Nuggets with Fresh Aloe vera Gel

    PubMed Central

    Rajkumar, V.; Verma, Arun K.; Patra, G.; Pradhan, S.; Biswas, S.; Chauhan, P.; Das, Arun K.

    2016-01-01

    Aloe vera has been used worldwide for pharmaceutical, food, and cosmetic industries due to its wide biological activities. However, quality improvement of low fat meat products and their acceptability with added Aloe vera gel (AVG) is scanty. The aim of this study was to explore the feasibility of using fresh AVG on physicochemical, textural, sensory and nutritive qualities of goat meat nuggets. The products were prepared with 0%, 2.5%, and 5% fresh AVG replacing goat meat and were analyzed for proximate composition, physicochemical and textural properties, fatty acid profile and sensory parameters. Changes in lipid oxidation and microbial growth of nuggets were also evaluated over 9 days of refrigerated storage. The results showed that AVG significantly (p<0.05) decreased the pH value and protein content of meat emulsion and nuggets. Product yield was affected at 5% level of gel. Addition of AVG in the formulation significantly affected the values of texture profile analysis. The AVG reduced the lipid oxidation and microbial growth in nuggets during storage. Sensory panelists preferred nuggets with 2.5% AVG over nuggets with 5% AVG. Therefore, AVG up to 2.5% level could be used for quality improvement in goat meat nuggets without affecting its sensorial, textural and nutritive values. PMID:26954177

  17. Predicting perceptual quality of images in realistic scenario using deep filter banks

    NASA Astrophysics Data System (ADS)

    Zhang, Weixia; Yan, Jia; Hu, Shiyong; Ma, Yang; Deng, Dexiang

    2018-03-01

    Classical image perceptual quality assessment models usually resort to natural scene statistic methods, which are based on an assumption that certain reliable statistical regularities hold on undistorted images and will be corrupted by introduced distortions. However, these models usually fail to accurately predict degradation severity of images in realistic scenarios since complex, multiple, and interactive authentic distortions usually appear on them. We propose a quality prediction model based on convolutional neural network. Quality-aware features extracted from filter banks of multiple convolutional layers are aggregated into the image representation. Furthermore, an easy-to-implement and effective feature selection strategy is used to further refine the image representation and finally a linear support vector regression model is trained to map image representation into images' subjective perceptual quality scores. The experimental results on benchmark databases present the effectiveness and generalizability of the proposed model.

  18. Image quality assessment by preprocessing and full reference model combination

    NASA Astrophysics Data System (ADS)

    Bianco, S.; Ciocca, G.; Marini, F.; Schettini, R.

    2009-01-01

    This paper focuses on full-reference image quality assessment and presents different computational strategies aimed to improve the robustness and accuracy of some well known and widely used state of the art models, namely the Structural Similarity approach (SSIM) by Wang and Bovik and the S-CIELAB spatial-color model by Zhang and Wandell. We investigate the hypothesis that combining error images with a visual attention model could allow a better fit of the psycho-visual data of the LIVE Image Quality assessment Database Release 2. We show that the proposed quality assessment metric better correlates with the experimental data.

  19. PLÉIADES Project: Assessment of Georeferencing Accuracy, Image Quality, Pansharpening Performence and Dsm/dtm Quality

    NASA Astrophysics Data System (ADS)

    Topan, Hüseyin; Cam, Ali; Özendi, Mustafa; Oruç, Murat; Jacobsen, Karsten; Taşkanat, Talha

    2016-06-01

    Pléiades 1A and 1B are twin optical satellites of Optical and Radar Federated Earth Observation (ORFEO) program jointly running by France and Italy. They are the first satellites of Europe with sub-meter resolution. Airbus DS (formerly Astrium Geo) runs a MyGIC (formerly Pléiades Users Group) program to validate Pléiades images worldwide for various application purposes. The authors conduct three projects, one is within this program, the second is supported by BEU Scientific Research Project Program, and the third is supported by TÜBİTAK. Assessment of georeferencing accuracy, image quality, pansharpening performance and Digital Surface Model/Digital Terrain Model (DSM/DTM) quality subjects are investigated in these projects. For these purposes, triplet panchromatic (50 cm Ground Sampling Distance (GSD)) and VNIR (2 m GSD) Pléiades 1A images were investigated over Zonguldak test site (Turkey) which is urbanised, mountainous and covered by dense forest. The georeferencing accuracy was estimated with a standard deviation in X and Y (SX, SY) in the range of 0.45m by bias corrected Rational Polynomial Coefficient (RPC) orientation, using ~170 Ground Control Points (GCPs). 3D standard deviation of ±0.44m in X, ±0.51m in Y, and ±1.82m in Z directions have been reached in spite of the very narrow angle of convergence by bias corrected RPC orientation. The image quality was also investigated with respect to effective resolution, Signal to Noise Ratio (SNR) and blur coefficient. The effective resolution was estimated with factor slightly below 1.0, meaning that the image quality corresponds to the nominal resolution of 50cm. The blur coefficients were achieved between 0.39-0.46 for triplet panchromatic images, indicating a satisfying image quality. SNR is in the range of other comparable space borne images which may be caused by de-noising of Pléiades images. The pansharpened images were generated by various methods, and are validated by most common statistical

  20. 40 CFR 89.510 - Compliance with acceptable quality level and passing and failing criteria for selective...

    Code of Federal Regulations, 2011 CFR

    2011-07-01

    ... 40 Protection of Environment 20 2011-07-01 2011-07-01 false Compliance with acceptable quality level and passing and failing criteria for selective enforcement audits. 89.510 Section 89.510... EMISSIONS FROM NEW AND IN-USE NONROAD COMPRESSION-IGNITION ENGINES Selective Enforcement Auditing § 89.510...

  1. Image quality and stability of image-guided radiotherapy (IGRT) devices: A comparative study.

    PubMed

    Stock, Markus; Pasler, Marlies; Birkfellner, Wolfgang; Homolka, Peter; Poetter, Richard; Georg, Dietmar

    2009-10-01

    Our aim was to implement standards for quality assurance of IGRT devices used in our department and to compare their performances with that of a CT simulator. We investigated image quality parameters for three devices over a period of 16months. A multislice CT was used as a benchmark and results related to noise, spatial resolution, low contrast visibility (LCV) and uniformity were compared with a cone beam CT (CBCT) at a linac and simulator. All devices performed well in terms of LCV and, in fact, exceeded vendor specifications. MTF was comparable between CT and linac CBCT. Integral nonuniformity was, on average, 0.002 for the CT and 0.006 for the linac CBCT. Uniformity, LCV and MTF varied depending on the protocols used for the linac CBCT. Contrast-to-noise ratio was an average of 51% higher for the CT than for the linac and simulator CBCT. No significant time trend was observed and tolerance limits were implemented. Reasonable differences in image quality between CT and CBCT were observed. Further research and development are necessary to increase image quality of commercially available CBCT devices in order for them to serve the needs for adaptive and/or online planning.

  2. Locally Enhanced Image Quality with Tunable Hybrid Metasurfaces

    NASA Astrophysics Data System (ADS)

    Shchelokova, Alena V.; Slobozhanyuk, Alexey P.; Melchakova, Irina V.; Glybovski, Stanislav B.; Webb, Andrew G.; Kivshar, Yuri S.; Belov, Pavel A.

    2018-01-01

    Metasurfaces represent a new paradigm in artificial subwavelength structures due to their potential to overcome many challenges typically associated with bulk metamaterials. The ability to make very thin structures and change their properties dynamically makes metasurfaces an exceptional meta-optics platform for engineering advanced electromagnetic and photonic metadevices. Here, we suggest and demonstrate experimentally a tunable metasurface capable of enhancing significantly the local image quality in magnetic resonance imaging. We present a design of the hybrid metasurface based on electromagnetically coupled dielectric and metallic elements. We demonstrate how to tailor the spectral characteristics of the metasurface eigenmodes by changing dynamically the effective permittivity of the structure. By maximizing a coupling between metasurface eigenmodes and transmitted and received fields in the magnetic resonance imaging (MRI) system, we enhance the device sensitivity that results in a substantial improvement of the image quality.

  3. Prostate seed implant quality assessment using MR and CT image fusion.

    PubMed

    Amdur, R J; Gladstone, D; Leopold, K A; Harris, R D

    1999-01-01

    After a seed implant of the prostate, computerized tomography (CT) is ideal for determining seed distribution but soft tissue anatomy is frequently not well visualized. Magnetic resonance (MR) images soft tissue anatomy well but seed visualization is problematic. We describe a method of fusing CT and MR images to exploit the advantages of both of these modalities when assessing the quality of a prostate seed implant. Eleven consecutive prostate seed implant patients were imaged with axial MR and CT scans. MR and CT images were fused in three dimensions using the Pinnacle 3.0 version of the ADAC treatment planning system. The urethra and bladder base were used to "line up" MR and CT image sets during image fusion. Alignment was accomplished using translation and rotation in the three ortho-normal planes. Accuracy of image fusion was evaluated by calculating the maximum deviation in millimeters between the center of the urethra on axial MR versus CT images. Implant quality was determined by comparing dosimetric results to previously set parameters. Image fusion was performed with a high degree of accuracy. When lining up the urethra and base of bladder, the maximum difference in axial position of the urethra between MR and CT averaged 2.5 mm (range 1.3-4.0 mm, SD 0.9 mm). By projecting CT-derived dose distributions over MR images of soft tissue structures, qualitative and quantitative evaluation of implant quality is straightforward. The image-fusion process we describe provides a sophisticated way of assessing the quality of a prostate seed implant. Commercial software makes the process time-efficient and available to any clinical practice with a high-quality treatment planning system. While we use MR to image soft tissue structures, the process could be used with any imaging modality that is able to visualize the prostatic urethra (e.g., ultrasound).

  4. Imaging study of using radiopharmaceuticals labeled with cyclotron-produced 99mTc.

    PubMed

    Hou, X; Tanguay, J; Vuckovic, M; Buckley, K; Schaffer, P; Bénard, F; Ruth, T J; Celler, A

    2016-12-07

    Cyclotron-produced 99m Tc (CPTc) has been recognized as an attractive and practical substitution of reactor/generator based 99m Tc. However, the small amount of 92-98 Mo in the irradiation of enriched 100 Mo could lead to the production of other radioactive technetium isotopes (Tc-impurities) which cannot be chemically separated. Thus, these impurities could contribute to patient dose and affect image quality. The potential radiation dose caused by these Tc-impurities produced using different targets, irradiation conditions, and corresponding to different injection times have been investigated, leading us to create dose-based limits of these parameters for producing clinically acceptable CPTc. However, image quality has been not considered. The aim of the present work is to provide a comprehensive and quantitative analysis of image quality for CPTc. The impact of Tc-impurities in CPTc on image resolution, background noise, and contrast is investigated by performing both Monte-Carlo simulations and phantom experiments. Various targets, irradiation, and acquisition conditions are employed for investigating the image-based limits of CPTc production parameters. Additionally, the relationship between patient dose and image quality of CPTc samples is studied. Only those samples which meet both dose- and image-based limits should be accepted in future clinical studies.

  5. Imaging study of using radiopharmaceuticals labeled with cyclotron-produced 99mTc

    NASA Astrophysics Data System (ADS)

    Hou, X.; Tanguay, J.; Vuckovic, M.; Buckley, K.; Schaffer, P.; Bénard, F.; Ruth, T. J.; Celler, A.

    2016-12-01

    Cyclotron-produced 99mTc (CPTc) has been recognized as an attractive and practical substitution of reactor/generator based 99mTc. However, the small amount of 92-98Mo in the irradiation of enriched 100Mo could lead to the production of other radioactive technetium isotopes (Tc-impurities) which cannot be chemically separated. Thus, these impurities could contribute to patient dose and affect image quality. The potential radiation dose caused by these Tc-impurities produced using different targets, irradiation conditions, and corresponding to different injection times have been investigated, leading us to create dose-based limits of these parameters for producing clinically acceptable CPTc. However, image quality has been not considered. The aim of the present work is to provide a comprehensive and quantitative analysis of image quality for CPTc. The impact of Tc-impurities in CPTc on image resolution, background noise, and contrast is investigated by performing both Monte-Carlo simulations and phantom experiments. Various targets, irradiation, and acquisition conditions are employed for investigating the image-based limits of CPTc production parameters. Additionally, the relationship between patient dose and image quality of CPTc samples is studied. Only those samples which meet both dose- and image-based limits should be accepted in future clinical studies.

  6. TU-AB-207A-03: Image Quality, Dose, and Clinical Applications

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Dong, F.

    Practicing medical physicists are often time charged with the tasks of evaluating and troubleshooting complex image quality issues related to CT scanners. This course will equip them with a solid and practical understanding of common CT imaging chain and its major components with emphasis on acquisition physics and hardware, reconstruction, artifacts, image quality, dose, and advanced clinical applications. The core objective is to explain the effects of these major system components on the image quality. This course will not focus on the rapid-changing advanced technologies given the two-hour time limit, but the fundamental principles discussed in this course may facilitatemore » better understanding of those more complicated technologies. The course will begin with an overview of CT acquisition physics and geometry. X-ray tube and CT detector are important acquisition hardware critical to the overall image quality. Each of these two subsystems consists of several major components. An in-depth description of the function and failure modes of these components will be provided. Examples of artifacts related to these failure modes will be presented: off-focal radiation, tube arcing, heel effect, oil bubble, offset drift effect, cross-talk effect, and bad pixels. The fundamentals of CT image reconstruction will first be discussed on an intuitive level. Approaches that do not require rigorous derivation of mathematical formulations will be presented. This is followed by a detailed derivation of the Fourier slice theorem: the foundation of the FBP algorithm. FBP for parallel-beam, fan-beam, and cone-beam geometries will be discussed. To address the issue of radiation dose related to x-ray CT, recent advances in iterative reconstruction, their advantages, and clinical applications will also be described. Because of the nature of fundamental physics and mathematics, limitations in data acquisition, and non-ideal conditions of major system components, image artifact

  7. 48 CFR 46.502 - Responsibility for acceptance.

    Code of Federal Regulations, 2010 CFR

    2010-10-01

    ... 48 Federal Acquisition Regulations System 1 2010-10-01 2010-10-01 false Responsibility for acceptance. 46.502 Section 46.502 Federal Acquisition Regulations System FEDERAL ACQUISITION REGULATION CONTRACT MANAGEMENT QUALITY ASSURANCE Acceptance 46.502 Responsibility for acceptance. Acceptance of...

  8. Development of ultrasound/endoscopy PACS (picture archiving and communication system) and investigation of compression method for cine images

    NASA Astrophysics Data System (ADS)

    Osada, Masakazu; Tsukui, Hideki

    2002-09-01

    ABSTRACT Picture Archiving and Communication System (PACS) is a system which connects imaging modalities, image archives, and image workstations to reduce film handling cost and improve hospital workflow. Handling diagnostic ultrasound and endoscopy images is challenging, because it produces large amount of data such as motion (cine) images of 30 frames per second, 640 x 480 in resolution, with 24-bit color. Also, it requires enough image quality for clinical review. We have developed PACS which is able to manage ultrasound and endoscopy cine images with above resolution and frame rate, and investigate suitable compression method and compression rate for clinical image review. Results show that clinicians require capability for frame-by-frame forward and backward review of cine images because they carefully look through motion images to find certain color patterns which may appear in one frame. In order to satisfy this quality, we have chosen motion JPEG, installed and confirmed that we could capture this specific pattern. As for acceptable image compression rate, we have performed subjective evaluation. No subjects could tell the difference between original non-compressed images and 1:10 lossy compressed JPEG images. One subject could tell the difference between original and 1:20 lossy compressed JPEG images although it is acceptable. Thus, ratios of 1:10 to 1:20 are acceptable to reduce data amount and cost while maintaining quality for clinical review.

  9. 40 CFR 90.510 - Compliance with acceptable quality level and passing and failing criteria for selective...

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... level and passing and failing criteria for selective enforcement audits. 90.510 Section 90.510....510 Compliance with acceptable quality level and passing and failing criteria for selective... failed engine is an engine whose final test results pursuant to § 90.509(b), for one or more of the...

  10. 40 CFR 90.510 - Compliance with acceptable quality level and passing and failing criteria for selective...

    Code of Federal Regulations, 2013 CFR

    2013-07-01

    ... level and passing and failing criteria for selective enforcement audits. 90.510 Section 90.510....510 Compliance with acceptable quality level and passing and failing criteria for selective... failed engine is an engine whose final test results pursuant to § 90.509(b), for one or more of the...

  11. The medium and the message: a revisionist view of image quality

    NASA Astrophysics Data System (ADS)

    Ferwerda, James A.

    2010-02-01

    In his book "Understanding Media" social theorist Marshall McLuhan declared: "The medium is the message." The thesis of this paper is that with respect to image quality, imaging system developers have taken McLuhan's dictum too much to heart. Efforts focus on improving the technical specifications of the media (e.g. dynamic range, color gamut, resolution, temporal response) with little regard for the visual messages the media will be used to communicate. We present a series of psychophysical studies that investigate the visual system's ability to "see through" the limitations of imaging media to perceive the messages (object and scene properties) the images represent. The purpose of these studies is to understand the relationships between the signal characteristics of an image and the fidelity of the visual information the image conveys. The results of these studies provide a new perspective on image quality that shows that images that may be very different in "quality", can be visually equivalent as realistic representations of objects and scenes.

  12. An image quality comparison study between XVI and OBI CBCT systems.

    PubMed

    Kamath, Srijit; Song, William; Chvetsov, Alexei; Ozawa, Shuichi; Lu, Haibin; Samant, Sanjiv; Liu, Chihray; Li, Jonathan G; Palta, Jatinder R

    2011-02-04

    The purpose of this study is to evaluate and compare image quality characteristics for two commonly used and commercially available CBCT systems: the X-ray Volumetric Imager and the On-Board Imager. A commonly used CATPHAN image quality phantom was used to measure various image quality parameters, namely, pixel value stability and accuracy, noise, contrast to noise ratio (CNR), high-contrast resolution, low contrast resolution and image uniformity. For the XVI unit, we evaluated the image quality for four manufacturer-supplied protocols as a function of mAs. For the OBI unit, we did the same for the full-fan and half-fan scanning modes, which were respectively used with the full bow-tie and half bow-tie filters. For XVI, the mean pixel values of regions of interest were found to generally decrease with increasing mAs for all protocols, while they were relatively stable with mAs for OBI. Noise was slightly lower on XVI and was seen to decrease with increasing mAs, while CNR increased with mAs for both systems. For XVI and OBI, the high-contrast resolution was approximately limited by the pixel resolution of the reconstructed image. On OBI images, up to 6 and 5 discs of 1% and 0.5% contrast, respectively, were visible for a high mAs setting using the full-fan mode, while none of the discs were clearly visible on the XVI images for various mAs settings when the medium resolution reconstruction was used. In conclusion, image quality parameters for XVI and OBI have been quantified and compared for clinical protocols under various mAs settings. These results need to be viewed in the context of a recent study that reported the dose-mAs relationship for the two systems and found that OBI generally delivered higher imaging doses than XVI.

  13. FFDM image quality assessment using computerized image texture analysis

    NASA Astrophysics Data System (ADS)

    Berger, Rachelle; Carton, Ann-Katherine; Maidment, Andrew D. A.; Kontos, Despina

    2010-04-01

    Quantitative measures of image quality (IQ) are routinely obtained during the evaluation of imaging systems. These measures, however, do not necessarily correlate with the IQ of the actual clinical images, which can also be affected by factors such as patient positioning. No quantitative method currently exists to evaluate clinical IQ. Therefore, we investigated the potential of using computerized image texture analysis to quantitatively assess IQ. Our hypothesis is that image texture features can be used to assess IQ as a measure of the image signal-to-noise ratio (SNR). To test feasibility, the "Rachel" anthropomorphic breast phantom (Model 169, Gammex RMI) was imaged with a Senographe 2000D FFDM system (GE Healthcare) using 220 unique exposure settings (target/filter, kVs, and mAs combinations). The mAs were varied from 10%-300% of that required for an average glandular dose (AGD) of 1.8 mGy. A 2.5cm2 retroareolar region of interest (ROI) was segmented from each image. The SNR was computed from the ROIs segmented from images linear with dose (i.e., raw images) after flat-field and off-set correction. Image texture features of skewness, coarseness, contrast, energy, homogeneity, and fractal dimension were computed from the Premium ViewTM postprocessed image ROIs. Multiple linear regression demonstrated a strong association between the computed image texture features and SNR (R2=0.92, p<=0.001). When including kV, target and filter as additional predictor variables, a stronger association with SNR was observed (R2=0.95, p<=0.001). The strong associations indicate that computerized image texture analysis can be used to measure image SNR and potentially aid in automating IQ assessment as a component of the clinical workflow. Further work is underway to validate our findings in larger clinical datasets.

  14. CLINICAL AUDIT OF IMAGE QUALITY IN RADIOLOGY USING VISUAL GRADING CHARACTERISTICS ANALYSIS.

    PubMed

    Tesselaar, Erik; Dahlström, Nils; Sandborg, Michael

    2016-06-01

    The aim of this work was to assess whether an audit of clinical image quality could be efficiently implemented within a limited time frame using visual grading characteristics (VGC) analysis. Lumbar spine radiography, bedside chest radiography and abdominal CT were selected. For each examination, images were acquired or reconstructed in two ways. Twenty images per examination were assessed by 40 radiology residents using visual grading of image criteria. The results were analysed using VGC. Inter-observer reliability was assessed. The results of the visual grading analysis were consistent with expected outcomes. The inter-observer reliability was moderate to good and correlated with perceived image quality (r(2) = 0.47). The median observation time per image or image series was within 2 min. These results suggest that the use of visual grading of image criteria to assess the quality of radiographs provides a rapid method for performing an image quality audit in a clinical environment. © The Author 2015. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  15. Image quality assessment of silent T2 PROPELLER sequence for brain imaging in infants.

    PubMed

    Kim, Hyun Gi; Choi, Jin Wook; Yoon, Soo Han; Lee, Sieun

    2018-02-01

    Infants are vulnerable to high acoustic noise. Acoustic noise generated by MR scanning can be reduced by a silent sequence. The purpose of this study is to compare the image quality of the conventional and silent T2 PROPELLER sequences for brain imaging in infants. A total of 36 scans were acquired from 24 infants using a 3 T MR scanner. Each patient underwent both conventional and silent T2 PROPELLER sequences. Acoustic noise level was measured. Quantitative and qualitative assessments were performed with the images taken with each sequence. The sound pressure level of the conventional T2 PROPELLER imaging sequence was 92.1 dB and that of the silent T2 PROPELLER imaging sequence was 73.3 dB (reduction of 20%). On quantitative assessment, the two sequences (conventional vs silent T2 PROPELLER) did not show significant difference in relative contrast (0.069 vs 0.068, p value = 0.536) and signal-to-noise ratio (75.4 vs 114.8, p value = 0.098). Qualitative assessment of overall image quality (p value = 0.572), grey-white differentiation (p value = 0.986), shunt-related artefact (p value > 0.999), motion artefact (p value = 0.801) and myelination degree in different brain regions (p values ≥ 0.092) did not show significant difference between the two sequences. The silent T2 PROPELLER sequence reduces acoustic noise and generated comparable image quality to that of the conventional sequence. Advances in knowledge: This is the first report to compare silent T2 PROPELLER images with that of conventional T2 PROPELLER images in children.

  16. SU-E-J-36: Comparison of CBCT Image Quality for Manufacturer Default Imaging Modes

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Nelson, G

    Purpose CBCT is being increasingly used in patient setup for radiotherapy. Often the manufacturer default scan modes are used for performing these CBCT scans with the assumption that they are the best options. To quantitatively assess the image quality of these scan modes, all of the scan modes were tested as well as options with the reconstruction algorithm. Methods A CatPhan 504 phantom was scanned on a TrueBeam Linear Accelerator using the manufacturer scan modes (FSRT Head, Head, Image Gently, Pelvis, Pelvis Obese, Spotlight, & Thorax). The Head mode scan was then reconstructed multiple times with all filter options (Smooth,more » Standard, Sharp, & Ultra Sharp) and all Ring Suppression options (Disabled, Weak, Medium, & Strong). An open source ImageJ tool was created for analyzing the CatPhan 504 images. Results The MTF curve was primarily dictated by the voxel size and the filter used in the reconstruction algorithm. The filters also impact the image noise. The CNR was worst for the Image Gently mode, followed by FSRT Head and Head. The sharper the filter, the worse the CNR. HU varied significantly between scan modes. Pelvis Obese had lower than expected HU values than most while the Image Gently mode had higher than expected HU values. If a therapist tried to use preset window and level settings, they would not show the desired tissue for some scan modes. Conclusion Knowing the image quality of the set scan modes, will enable users to better optimize their setup CBCT. Evaluation of the scan mode image quality could improve setup efficiency and lead to better treatment outcomes.« less

  17. Evaluation of image quality in terahertz pulsed imaging using test objects.

    PubMed

    Fitzgerald, A J; Berry, E; Miles, R E; Zinovev, N N; Smith, M A; Chamberlain, J M

    2002-11-07

    As with other imaging modalities, the performance of terahertz (THz) imaging systems is limited by factors of spatial resolution, contrast and noise. The purpose of this paper is to introduce test objects and image analysis methods to evaluate and compare THz image quality in a quantitative and objective way, so that alternative terahertz imaging system configurations and acquisition techniques can be compared, and the range of image parameters can be assessed. Two test objects were designed and manufactured, one to determine the modulation transfer functions (MTF) and the other to derive image signal to noise ratio (SNR) at a range of contrasts. As expected the higher THz frequencies had larger MTFs, and better spatial resolution as determined by the spatial frequency at which the MTF dropped below the 20% threshold. Image SNR was compared for time domain and frequency domain image parameters and time delay based images consistently demonstrated higher SNR than intensity based parameters such as relative transmittance because the latter are more strongly affected by the sources of noise in the THz system such as laser fluctuations and detector shot noise.

  18. Learning a No-Reference Quality Assessment Model of Enhanced Images With Big Data.

    PubMed

    Gu, Ke; Tao, Dacheng; Qiao, Jun-Fei; Lin, Weisi

    2018-04-01

    In this paper, we investigate into the problem of image quality assessment (IQA) and enhancement via machine learning. This issue has long attracted a wide range of attention in computational intelligence and image processing communities, since, for many practical applications, e.g., object detection and recognition, raw images are usually needed to be appropriately enhanced to raise the visual quality (e.g., visibility and contrast). In fact, proper enhancement can noticeably improve the quality of input images, even better than originally captured images, which are generally thought to be of the best quality. In this paper, we present two most important contributions. The first contribution is to develop a new no-reference (NR) IQA model. Given an image, our quality measure first extracts 17 features through analysis of contrast, sharpness, brightness and more, and then yields a measure of visual quality using a regression module, which is learned with big-data training samples that are much bigger than the size of relevant image data sets. The results of experiments on nine data sets validate the superiority and efficiency of our blind metric compared with typical state-of-the-art full-reference, reduced-reference and NA IQA methods. The second contribution is that a robust image enhancement framework is established based on quality optimization. For an input image, by the guidance of the proposed NR-IQA measure, we conduct histogram modification to successively rectify image brightness and contrast to a proper level. Thorough tests demonstrate that our framework can well enhance natural images, low-contrast images, low-light images, and dehazed images. The source code will be released at https://sites.google.com/site/guke198701/publications.

  19. Don’t forget the posters! Quality and content variables associated with accepted abstracts at a national trauma meeting

    PubMed Central

    Dossett, Lesly A.; Fox, Erin E.; del Junco, Deborah J.; Zaydfudim, Victor; Kauffmann, Rondi; Shelton, Julia; Wang, Weiwei; Cioffi, William G.; Holcomb, John B.; Cotton, Bryan A.

    2014-01-01

    BACKGROUND As a primary venue for presenting research results, abstracts selected for presentation at national meetings should be of the highest scientific merit and research quality. It is uncertain to what degree this is achieved as the methodological quality of abstracts submitted to national surgical meetings has not been previously described. The objective of this study was to evaluate abstracts presented at a leading trauma meeting for methodological quality. METHODS All abstracts accepted for the 2009 American Association for the Surgery of Trauma meeting were reviewed and scored for methodological quality based on 10 criteria (scores, 0–10; 10 being the highest). Criteria were based on nationally published methodology guidelines. Two independent reviewers who were blinded to institution, region, and author reviewed each abstract. RESULTS A total of 187 abstracts were accepted for presentation (67 oral and 120 posters). The most frequent clinical topics were shock/transfusion (23%), abdomen (12%), and nervous system (11%). Shock/transfusion abstracts were more common in the oral presentations (31% vs. 19%; p =0.06). Abstracts from the northeast and south regions were the most common in both oral (26% and 29%) and posters (25% and 24%). Basic science accounted for 12% of accepted studies, while 51% were clinical and 28% were health services/outcomes. Only 8% of abstracts presented randomized data and only 11% reported null findings. Overall abstract scores ranged from 3 to 10 (median, 7; mean, 7.4). Abstracts selected for poster presentation had an overall higher score than those selected for oral presentation (7.4 ±1.7 vs. 6.8 ±1.7; p =0.02). CONCLUSION Although oral presentations traditionally receive the most attention and interest, the methodological quality of abstracts accepted for poster presentation equals (and sometimes exceeds) that of oral abstracts. Attendees of these national meetings should reconsider their time spent in viewing and visiting

  20. Assessing the quality of restored images in optical long-baseline interferometry

    NASA Astrophysics Data System (ADS)

    Gomes, Nuno; Garcia, Paulo J. V.; Thiébaut, Éric

    2017-03-01

    Assessing the quality of aperture synthesis maps is relevant for benchmarking image reconstruction algorithms, for the scientific exploitation of data from optical long-baseline interferometers, and for the design/upgrade of new/existing interferometric imaging facilities. Although metrics have been proposed in these contexts, no systematic study has been conducted on the selection of a robust metric for quality assessment. This article addresses the question: what is the best metric to assess the quality of a reconstructed image? It starts by considering several metrics and selecting a few based on general properties. Then, a variety of image reconstruction cases are considered. The observational scenarios are phase closure and phase referencing at the Very Large Telescope Interferometer (VLTI), for a combination of two, three, four and six telescopes. End-to-end image reconstruction is accomplished with the MIRA software, and several merit functions are put to test. It is found that convolution by an effective point spread function is required for proper image quality assessment. The effective angular resolution of the images is superior to naive expectation based on the maximum frequency sampled by the array. This is due to the prior information used in the aperture synthesis algorithm and to the nature of the objects considered. The ℓ1-norm is the most robust of all considered metrics, because being linear it is less sensitive to image smoothing by high regularization levels. For the cases considered, this metric allows the implementation of automatic quality assessment of reconstructed images, with a performance similar to human selection.

  1. The relationship between resident burnout and safety-related and acceptability-related quality of healthcare: a systematic literature review.

    PubMed

    Dewa, Carolyn S; Loong, Desmond; Bonato, Sarah; Trojanowski, Lucy; Rea, Margaret

    2017-11-09

    There has been increasing interest in examining the relationship between physician wellbeing and quality of patient care. However, few reviews have specifically focused on resident burnout and quality of patient care. The purpose of this systematic literature review of the current scientific literature is to address the question, "How does resident burnout affect the quality of healthcare related to the dimensions of acceptability and safety?" This systematic literature review uses a multi-step screening process of publicly available peer-reviewed studies from five electronic databases: (1) Medline Current, (2) Medline In-process, (3) PsycINFO, (4) Embase, and (5) Web of Science. The electronic literature search resulted in the identification of 4638 unique citations. Of these, 10 articles were included in the review. Studies were assessed for risk of bias. Of the 10 studies that met the inclusion criteria, eight were conducted in the US, one in The Netherlands, and one in Mexico. Eight of the 10 studies focused on patient safety. The results of these included studies suggest there is moderate evidence that burnout is associated with patient safety (i.e., resident self-perceived medical errors and sub-optimal care). There is less evidence that specific dimensions of burnout are related to acceptability (i.e., quality of care, communication with patients). The results of this systematic literature review suggest a relationship between patient safety and burnout. These results potentially have important implications for the medical training milieu because residents are still in training and at the same time are asked to teach students. The results also indicate a need for more evidence-based interventions that support continued research examining quality of care measures, especially as they relate to acceptability.

  2. Optimisation of radiation dose and image quality in mobile neonatal chest radiography.

    PubMed

    Hinojos-Armendáriz, V I; Mejía-Rosales, S J; Franco-Cabrera, M C

    2018-05-01

    To optimise the radiation dose and image quality for chest radiography in the neonatal intensive care unit (NICU) by increasing the mean beam energy. Two techniques for the acquisition of NICU AP chest X-ray images were compared for image quality and radiation dose. 73 images were acquired using a standard technique (56 kV, 3.2 mAs and no additional filtration) and 90 images with a new technique (62 kV, 2 mAs and 2 mm Al filtration). The entrance surface air kerma (ESAK) was measured using a phantom and compared between the techniques and against established diagnostic reference levels (DRL). Images were evaluated using seven image quality criteria independently by three radiologists. Images quality and radiation dose were compared statistically between the standard and new techniques. The maximum ESAK for the new technique was 40.20 μGy, 43.7% of the ESAK of the standard technique. Statistical evaluation demonstrated no significant differences in image quality between the two acquisition techniques. Based on the techniques and acquisition factors investigated within this study, it is possible to lower the radiation dose without any significant effects on image quality by adding filtration (2 mm Al) and increasing the tube potential. Such steps are relatively simple to undertake and as such, other departments should consider testing and implementing this dose reduction strategy within clinical practice where appropriate. Copyright © 2017 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.

  3. The role of acceptance and values in quality of life in patients with an acquired brain injury: a questionnaire study.

    PubMed

    Van Bost, Gunther; Van Damme, Stefaan; Crombez, Geert

    2017-01-01

    An acquired brain injury (ABI) is a challenge for an individual's quality of life (QOL). In several chronic illnesses acceptance has been found to be associated with a better health-related quality of life. This study investigated whether this relationship is also found in patients with ABI. We also explored the impact of the perceived ability to live according to one's own values (life-values-match). A total of 68 individuals (18-65 years of age) with an acquired brain injury completed a battery of questionnaires. The relations between health-related QOL (SF-36) and disease specific QOL (EBIQ; European Brain Injury Questionnaire), and personal values (Schwartz Values Inventory) and acceptance (ICQ; Illness Cognitions Questionnaire) were investigated. An additional question measured the life-values-match. Rehabilitation professionals reported the extent of impairment involved. Acceptance was positively associated with mental aspects of health-related QOL and the EBIQ Core Scale, after demographic variables and the extent of impairment were introduced in the regression. In a post hoc analysis we found that the life-values-match mediated the relationship between acceptance and mental aspects of QOL. In patients with an ABI, promoting acceptance may be useful to protect QOL. Strengthening the life-values-match may be a way to accomplish this.

  4. New patient-controlled abdominal compression method in radiography: radiation dose and image quality.

    PubMed

    Piippo-Huotari, Oili; Norrman, Eva; Anderzén-Carlsson, Agneta; Geijer, Håkan

    2018-05-01

    The radiation dose for patients can be reduced with many methods and one way is to use abdominal compression. In this study, the radiation dose and image quality for a new patient-controlled compression device were compared with conventional compression and compression in the prone position . To compare radiation dose and image quality of patient-controlled compression compared with conventional and prone compression in general radiography. An experimental design with quantitative approach. After obtaining the approval of the ethics committee, a consecutive sample of 48 patients was examined with the standard clinical urography protocol. The radiation doses were measured as dose-area product and analyzed with a paired t-test. The image quality was evaluated by visual grading analysis. Four radiologists evaluated each image individually by scoring nine criteria modified from the European quality criteria for diagnostic radiographic images. There was no significant difference in radiation dose or image quality between conventional and patient-controlled compression. Prone position resulted in both higher dose and inferior image quality. Patient-controlled compression gave similar dose levels as conventional compression and lower than prone compression. Image quality was similar with both patient-controlled and conventional compression and was judged to be better than in the prone position.

  5. PROPOSAL FOR A SIMPLE AND EFFICIENT MONTHLY QUALITY MANAGEMENT PROGRAM ASSESSING THE CONSISTENCY OF ROBOTIC IMAGE-GUIDED SMALL ANIMAL RADIATION SYSTEMS

    PubMed Central

    Brodin, N. Patrik; Guha, Chandan; Tomé, Wolfgang A.

    2015-01-01

    Modern pre-clinical radiation therapy (RT) research requires high precision and accurate dosimetry to facilitate the translation of research findings into clinical practice. Several systems are available that provide precise delivery and on-board imaging capabilities, highlighting the need for a quality management program (QMP) to ensure consistent and accurate radiation dose delivery. An ongoing, simple, and efficient QMP for image-guided robotic small animal irradiators used in pre-clinical RT research is described. Protocols were developed and implemented to assess the dose output constancy (based on the AAPM TG-61 protocol), cone-beam computed tomography (CBCT) image quality and object representation accuracy (using a custom-designed imaging phantom), CBCT-guided target localization accuracy and consistency of the CBCT-based dose calculation. To facilitate an efficient read-out and limit the user dependence of the QMP data analysis, a semi-automatic image analysis and data representation program was developed using the technical computing software MATLAB. The results of the first six months experience using the suggested QMP for a Small Animal Radiation Research Platform (SARRP) are presented, with data collected on a bi-monthly basis. The dosimetric output constancy was established to be within ±1 %, the consistency of the image resolution was within ±0.2 mm, the accuracy of CBCT-guided target localization was within ±0.5 mm, and dose calculation consistency was within ±2 s (± 3 %) per treatment beam. Based on these results, this simple quality assurance program allows for the detection of inconsistencies in dosimetric or imaging parameters that are beyond the acceptable variability for a reliable and accurate pre-clinical RT system, on a monthly or bi-monthly basis. PMID:26425981

  6. Proposal for a Simple and Efficient Monthly Quality Management Program Assessing the Consistency of Robotic Image-Guided Small Animal Radiation Systems.

    PubMed

    Brodin, N Patrik; Guha, Chandan; Tomé, Wolfgang A

    2015-11-01

    Modern pre-clinical radiation therapy (RT) research requires high precision and accurate dosimetry to facilitate the translation of research findings into clinical practice. Several systems are available that provide precise delivery and on-board imaging capabilities, highlighting the need for a quality management program (QMP) to ensure consistent and accurate radiation dose delivery. An ongoing, simple, and efficient QMP for image-guided robotic small animal irradiators used in pre-clinical RT research is described. Protocols were developed and implemented to assess the dose output constancy (based on the AAPM TG-61 protocol), cone-beam computed tomography (CBCT) image quality and object representation accuracy (using a custom-designed imaging phantom), CBCT-guided target localization accuracy and consistency of the CBCT-based dose calculation. To facilitate an efficient read-out and limit the user dependence of the QMP data analysis, a semi-automatic image analysis and data representation program was developed using the technical computing software MATLAB. The results of the first 6-mo experience using the suggested QMP for a Small Animal Radiation Research Platform (SARRP) are presented, with data collected on a bi-monthly basis. The dosimetric output constancy was established to be within ±1 %, the consistency of the image resolution was within ±0.2 mm, the accuracy of CBCT-guided target localization was within ±0.5 mm, and dose calculation consistency was within ±2 s (±3%) per treatment beam. Based on these results, this simple quality assurance program allows for the detection of inconsistencies in dosimetric or imaging parameters that are beyond the acceptable variability for a reliable and accurate pre-clinical RT system, on a monthly or bi-monthly basis.

  7. Beef quality labels: A combination of sensory acceptance test, stated willingness to pay, and choice-based conjoint analysis.

    PubMed

    Meyerding, Stephan G H; Gentz, Maria; Altmann, Brianne; Meier-Dinkel, Lisa

    2018-08-01

    Consumer perspectives of beef quality are complex, leading to a market that is increasingly differentiating. Thus, ongoing monitoring and assessment of changes in consumer perspectives is essential to identify changing market conditions. Often only credence and search characteristics are evaluated in consumer studies; therefore the object of the present study is to examine consumer preferences and perceptions towards beef steaks, also including experience characteristics, using a mixed methods approach. For this reason, 55 consumers participated in an experiment in Germany, including a sensory acceptance test, stated willingness to pay, and choice-based conjoint analysis (CBCA). Different quality characteristics were included, but a focus on the quality labels of 'dry aged beef', 'Block House beef', and 'Angus beef' was predominant throughout the experiment with the results showing that quality labels significantly increased overall liking as well as the stated willingness to pay. Quality labels were also the one of the most important characteristics in the conjoint analysis, after origin and price. The results of all applied methods are comparable for the characteristic quality label. The combination of sensory acceptance test and CBCA were additionally able to evaluate all three kinds of beef quality characteristics, which could not be evaluated together only using a single method. This suggests that a mixture of methods should be used to gain better knowledge on the true behavior of beef consumers. Experience and credence characteristics, including beef quality labels, present opportunities for future research as well as the potential for determining product and market differentiation. Copyright © 2018 Elsevier Ltd. All rights reserved.

  8. Image quality and stability of image-guided radiotherapy (IGRT) devices: A comparative study

    PubMed Central

    Stock, Markus; Pasler, Marlies; Birkfellner, Wolfgang; Homolka, Peter; Poetter, Richard; Georg, Dietmar

    2010-01-01

    Introduction Our aim was to implement standards for quality assurance of IGRT devices used in our department and to compare their performances with that of a CT simulator. Materials and methods We investigated image quality parameters for three devices over a period of 16 months. A multislice CT was used as a benchmark and results related to noise, spatial resolution, low contrast visibility (LCV) and uniformity were compared with a cone beam CT (CBCT) at a linac and simulator. Results All devices performed well in terms of LCV and, in fact, exceeded vendor specifications. MTF was comparable between CT and linac CBCT. Integral nonuniformity was, on average, 0.002 for the CT and 0.006 for the linac CBCT. Uniformity, LCV and MTF varied depending on the protocols used for the linac CBCT. Contrast-to-noise ratio was an average of 51% higher for the CT than for the linac and simulator CBCT. No significant time trend was observed and tolerance limits were implemented. Discussion Reasonable differences in image quality between CT and CBCT were observed. Further research and development are necessary to increase image quality of commercially available CBCT devices in order for them to serve the needs for adaptive and/or online planning. PMID:19695725

  9. Blind image quality assessment via probabilistic latent semantic analysis.

    PubMed

    Yang, Xichen; Sun, Quansen; Wang, Tianshu

    2016-01-01

    We propose a blind image quality assessment that is highly unsupervised and training free. The new method is based on the hypothesis that the effect caused by distortion can be expressed by certain latent characteristics. Combined with probabilistic latent semantic analysis, the latent characteristics can be discovered by applying a topic model over a visual word dictionary. Four distortion-affected features are extracted to form the visual words in the dictionary: (1) the block-based local histogram; (2) the block-based local mean value; (3) the mean value of contrast within a block; (4) the variance of contrast within a block. Based on the dictionary, the latent topics in the images can be discovered. The discrepancy between the frequency of the topics in an unfamiliar image and a large number of pristine images is applied to measure the image quality. Experimental results for four open databases show that the newly proposed method correlates well with human subjective judgments of diversely distorted images.

  10. T2- and diffusion-weighted magnetic resonance imaging at 3T for the detection of prostate cancer with and without endorectal coil: An intraindividual comparison of image quality and diagnostic performance.

    PubMed

    Baur, Alexander D J; Daqqaq, Tareef; Wagner, Moritz; Maxeiner, Andreas; Huppertz, Alexander; Renz, Diane; Hamm, Bernd; Fischer, Thomas; Durmus, Tahir

    2016-06-01

    To intraindividually compare image quality and diagnostic performance of multiparametric MRI (mpMRI) at 3T for the detection of prostate cancer (PCa) using a pelvic phased-array coil (PAC) and a combined endorectal and pelvic phased-array coil (ERC-PAC). Forty-five patients were prospectively included and received mpMRI of the prostate using a PAC and an ERC-PAC during one imaging session. Two radiologists evaluated image quality and the most suspicious lesion according to the PI-RADS scoring system. Results of MRI-TRUS-fusion biopsy of the prostate served as reference standard. Patient comfort and acceptance were assessed using a standardized questionnaire. Overall image quality for T2WI was rated significantly better with an ERC-PAC compared to a PAC (p=0.0038). The weighted kappa for PI-RADS scores for T2WI and DWI with a PAC and an ERC-PAC was 0.70 and 0.73, respectively. For a PI-RADS sum score including T2WI and DWI the area under the curve with a PAC and an ERC-PAC were 0.95-0.99 and 0.93-0.97, respectively (p=0.1395). For T2WI and DWI performed at 3T index PCa lesion identification and evaluation did not differ significantly with both coil setups. Patients preferred MRI without an ERC. Therefore, the use of an ERC may be omitted in a prostate cancer detection setting. Copyright © 2016 Elsevier Ireland Ltd. All rights reserved.

  11. In acceptance we trust? Conceptualising acceptance as a viable approach to NGO security management.

    PubMed

    Fast, Larissa A; Freeman, C Faith; O'Neill, Michael; Rowley, Elizabeth

    2013-04-01

    This paper documents current understanding of acceptance as a security management approach and explores issues and challenges non-governmental organisations (NGOs) confront when implementing an acceptance approach to security management. It argues that the failure of organisations to systematise and clearly articulate acceptance as a distinct security management approach and a lack of organisational policies and procedures concerning acceptance hinder its efficacy as a security management approach. The paper identifies key and cross-cutting components of acceptance that are critical to its effective implementation in order to advance a comprehensive and systematic concept of acceptance. The key components of acceptance illustrate how organisational and staff functions affect positively or negatively an organisation's acceptance, and include: an organisation's principles and mission, communications, negotiation, programming, relationships and networks, stakeholder and context analysis, staffing, and image. The paper contends that acceptance is linked not only to good programming, but also to overall organisational management and structures. © 2013 The Author(s). Journal compilation © Overseas Development Institute, 2013.

  12. Radiation dose and image quality in pediatric chest CT: effects of iterative reconstruction in normal weight and overweight children.

    PubMed

    Yoon, Haesung; Kim, Myung-Joon; Yoon, Choon-Sik; Choi, Jiin; Shin, Hyun Joo; Kim, Hyun Gi; Lee, Mi-Jung

    2015-03-01

    New CT reconstruction techniques may help reduce the burden of ionizing radiation. To quantify radiation dose reduction when performing pediatric chest CT using a low-dose protocol and 50% adaptive statistical iterative reconstruction (ASIR) compared with age/gender-matched chest CT using a conventional dose protocol and reconstructed with filtered back projection (control group) and to determine its effect on image quality in normal weight and overweight children. We retrospectively reviewed 40 pediatric chest CT (M:F = 21:19; range: 0.1-17 years) in both groups. Radiation dose was compared between the two groups using paired Student's t-test. Image quality including noise, sharpness, artifacts and diagnostic acceptability was subjectively assessed by three pediatric radiologists using a four-point scale (superior, average, suboptimal, unacceptable). Eight children in the ASIR group and seven in the control group were overweight. All radiation dose parameters were significantly lower in the ASIR group (P < 0.01) with a greater than 57% dose reduction in overweight children. Image noise was higher in the ASIR group in both normal weight and overweight children. Only one scan in the ASIR group (1/40, 2.5%) was rated as diagnostically suboptimal and there was no unacceptable study. In both normal weight and overweight children, the ASIR technique is associated with a greater than 57% mean dose reduction, without significantly impacting diagnostic image quality in pediatric chest CT examinations. However, CT scans in overweight children may have a greater noise level, even when using the ASIR technique.

  13. Functional magnetic resonance imaging of awake monkeys: some approaches for improving imaging quality

    PubMed Central

    Chen, Gang; Wang, Feng; Dillenburger, Barbara C.; Friedman, Robert M.; Chen, Li M.; Gore, John C.; Avison, Malcolm J.; Roe, Anna W.

    2011-01-01

    Functional magnetic resonance imaging (fMRI), at high magnetic field strength can suffer from serious degradation of image quality because of motion and physiological noise, as well as spatial distortions and signal losses due to susceptibility effects. Overcoming such limitations is essential for sensitive detection and reliable interpretation of fMRI data. These issues are particularly problematic in studies of awake animals. As part of our initial efforts to study functional brain activations in awake, behaving monkeys using fMRI at 4.7T, we have developed acquisition and analysis procedures to improve image quality with encouraging results. We evaluated the influence of two main variables on image quality. First, we show how important the level of behavioral training is for obtaining good data stability and high temporal signal-to-noise ratios. In initial sessions, our typical scan session lasted 1.5 hours, partitioned into short (<10 minutes) runs. During reward periods and breaks between runs, the monkey exhibited movements resulting in considerable image misregistrations. After a few months of extensive behavioral training, we were able to increase the length of individual runs and the total length of each session. The monkey learned to wait until the end of a block for fluid reward, resulting in longer periods of continuous acquisition. Each additional 60 training sessions extended the duration of each session by 60 minutes, culminating, after about 140 training sessions, in sessions that last about four hours. As a result, the average translational movement decreased from over 500 μm to less than 80 μm, a displacement close to that observed in anesthetized monkeys scanned in a 7 T horizontal scanner. Another major source of distortion at high fields arises from susceptibility variations. To reduce such artifacts, we used segmented gradient-echo echo-planar imaging (EPI) sequences. Increasing the number of segments significantly decreased susceptibility

  14. Fusion and quality analysis for remote sensing images using contourlet transform

    NASA Astrophysics Data System (ADS)

    Choi, Yoonsuk; Sharifahmadian, Ershad; Latifi, Shahram

    2013-05-01

    Recent developments in remote sensing technologies have provided various images with high spatial and spectral resolutions. However, multispectral images have low spatial resolution and panchromatic images have low spectral resolution. Therefore, image fusion techniques are necessary to improve the spatial resolution of spectral images by injecting spatial details of high-resolution panchromatic images. The objective of image fusion is to provide useful information by improving the spatial resolution and the spectral information of the original images. The fusion results can be utilized in various applications, such as military, medical imaging, and remote sensing. This paper addresses two issues in image fusion: i) image fusion method and ii) quality analysis of fusion results. First, a new contourlet-based image fusion method is presented, which is an improvement over the wavelet-based fusion. This fusion method is then applied to a case study to demonstrate its fusion performance. Fusion framework and scheme used in the study are discussed in detail. Second, quality analysis for the fusion results is discussed. We employed various quality metrics in order to analyze the fusion results both spatially and spectrally. Our results indicate that the proposed contourlet-based fusion method performs better than the conventional wavelet-based fusion methods.

  15. Application-Driven No-Reference Quality Assessment for Dermoscopy Images With Multiple Distortions.

    PubMed

    Xie, Fengying; Lu, Yanan; Bovik, Alan C; Jiang, Zhiguo; Meng, Rusong

    2016-06-01

    Dermoscopy images often suffer from blur and uneven illumination distortions that occur during acquisition, which can adversely influence consequent automatic image analysis results on potential lesion objects. The purpose of this paper is to deploy an algorithm that can automatically assess the quality of dermoscopy images. Such an algorithm could be used to direct image recapture or correction. We describe an application-driven no-reference image quality assessment (IQA) model for dermoscopy images affected by possibly multiple distortions. For this purpose, we created a multiple distortion dataset of dermoscopy images impaired by varying degrees of blur and uneven illumination. The basis of this model is two single distortion IQA metrics that are sensitive to blur and uneven illumination, respectively. The outputs of these two metrics are combined to predict the quality of multiply distorted dermoscopy images using a fuzzy neural network. Unlike traditional IQA algorithms, which use human subjective score as ground truth, here ground truth is driven by the application, and generated according to the degree of influence of the distortions on lesion analysis. The experimental results reveal that the proposed model delivers accurate and stable quality prediction results for dermoscopy images impaired by multiple distortions. The proposed model is effective for quality assessment of multiple distorted dermoscopy images. An application-driven concept for IQA is introduced, and at the same time, a solution framework for the IQA of multiple distortions is proposed.

  16. Image Acquisition and Quality in Digital Radiography.

    PubMed

    Alexander, Shannon

    2016-09-01

    Medical imaging has undergone dramatic changes and technological breakthroughs since the introduction of digital radiography. This article presents information on the development of digital radiography and types of digital radiography systems. Aspects of image quality and radiation exposure control are highlighted as well. In addition, the article includes related workplace changes and medicolegal considerations in the digital radiography environment. ©2016 American Society of Radiologic Technologists.

  17. WE-G-207-05: Relationship Between CT Image Quality, Segmentation Performance, and Quantitative Image Feature Analysis

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lee, J; Nishikawa, R; Reiser, I

    Purpose: Segmentation quality can affect quantitative image feature analysis. The objective of this study is to examine the relationship between computed tomography (CT) image quality, segmentation performance, and quantitative image feature analysis. Methods: A total of 90 pathology proven breast lesions in 87 dedicated breast CT images were considered. An iterative image reconstruction (IIR) algorithm was used to obtain CT images with different quality. With different combinations of 4 variables in the algorithm, this study obtained a total of 28 different qualities of CT images. Two imaging tasks/objectives were considered: 1) segmentation and 2) classification of the lesion as benignmore » or malignant. Twenty-three image features were extracted after segmentation using a semi-automated algorithm and 5 of them were selected via a feature selection technique. Logistic regression was trained and tested using leave-one-out-cross-validation and its area under the ROC curve (AUC) was recorded. The standard deviation of a homogeneous portion and the gradient of a parenchymal portion of an example breast were used as an estimate of image noise and sharpness. The DICE coefficient was computed using a radiologist’s drawing on the lesion. Mean DICE and AUC were used as performance metrics for each of the 28 reconstructions. The relationship between segmentation and classification performance under different reconstructions were compared. Distributions (median, 95% confidence interval) of DICE and AUC for each reconstruction were also compared. Results: Moderate correlation (Pearson’s rho = 0.43, p-value = 0.02) between DICE and AUC values was found. However, the variation between DICE and AUC values for each reconstruction increased as the image sharpness increased. There was a combination of IIR parameters that resulted in the best segmentation with the worst classification performance. Conclusion: There are certain images that yield better segmentation or

  18. Super-resolution convolutional neural network for the improvement of the image quality of magnified images in chest radiographs

    NASA Astrophysics Data System (ADS)

    Umehara, Kensuke; Ota, Junko; Ishimaru, Naoki; Ohno, Shunsuke; Okamoto, Kentaro; Suzuki, Takanori; Shirai, Naoki; Ishida, Takayuki

    2017-02-01

    Single image super-resolution (SR) method can generate a high-resolution (HR) image from a low-resolution (LR) image by enhancing image resolution. In medical imaging, HR images are expected to have a potential to provide a more accurate diagnosis with the practical application of HR displays. In recent years, the super-resolution convolutional neural network (SRCNN), which is one of the state-of-the-art deep learning based SR methods, has proposed in computer vision. In this study, we applied and evaluated the SRCNN scheme to improve the image quality of magnified images in chest radiographs. For evaluation, a total of 247 chest X-rays were sampled from the JSRT database. The 247 chest X-rays were divided into 93 training cases with non-nodules and 152 test cases with lung nodules. The SRCNN was trained using the training dataset. With the trained SRCNN, the HR image was reconstructed from the LR one. We compared the image quality of the SRCNN and conventional image interpolation methods, nearest neighbor, bilinear and bicubic interpolations. For quantitative evaluation, we measured two image quality metrics, peak signal-to-noise ratio (PSNR) and structural similarity (SSIM). In the SRCNN scheme, PSNR and SSIM were significantly higher than those of three interpolation methods (p<0.001). Visual assessment confirmed that the SRCNN produced much sharper edge than conventional interpolation methods without any obvious artifacts. These preliminary results indicate that the SRCNN scheme significantly outperforms conventional interpolation algorithms for enhancing image resolution and that the use of the SRCNN can yield substantial improvement of the image quality of magnified images in chest radiographs.

  19. A quality assurance framework for the fully automated and objective evaluation of image quality in cone-beam computed tomography.

    PubMed

    Steiding, Christian; Kolditz, Daniel; Kalender, Willi A

    2014-03-01

    Thousands of cone-beam computed tomography (CBCT) scanners for vascular, maxillofacial, neurological, and body imaging are in clinical use today, but there is no consensus on uniform acceptance and constancy testing for image quality (IQ) and dose yet. The authors developed a quality assurance (QA) framework for fully automated and time-efficient performance evaluation of these systems. In addition, the dependence of objective Fourier-based IQ metrics on direction and position in 3D volumes was investigated for CBCT. The authors designed a dedicated QA phantom 10 cm in length consisting of five compartments, each with a diameter of 10 cm, and an optional extension ring 16 cm in diameter. A homogeneous section of water-equivalent material allows measuring CT value accuracy, image noise and uniformity, and multidimensional global and local noise power spectra (NPS). For the quantitative determination of 3D high-contrast spatial resolution, the modulation transfer function (MTF) of centrally and peripherally positioned aluminum spheres was computed from edge profiles. Additional in-plane and axial resolution patterns were used to assess resolution qualitatively. The characterization of low-contrast detectability as well as CT value linearity and artifact behavior was tested by utilizing sections with soft-tissue-equivalent and metallic inserts. For an automated QA procedure, a phantom detection algorithm was implemented. All tests used in the dedicated QA program were initially verified in simulation studies and experimentally confirmed on a clinical dental CBCT system. The automated IQ evaluation of volume data sets of the dental CBCT system was achieved with the proposed phantom requiring only one scan for the determination of all desired parameters. Typically, less than 5 min were needed for phantom set-up, scanning, and data analysis. Quantitative evaluation of system performance over time by comparison to previous examinations was also verified. The maximum

  20. Body Image, Self-esteem, and Quality of Life in Patients with Psoriasis.

    PubMed

    Nazik, Hulya; Nazik, Selcuk; Gul, Feride C

    2017-01-01

    Psoriasis is a chronic inflammatory disease of the skin that may affect the visible areas of body. Hence, the quality of life, self-esteem, and body image can be affected in psoriasis patients. We aimed in the present study to assess the effects of psoriasis on the quality of life, self-esteem, and body image. The study included 92 patients with psoriasis, along with 98 control participants. The sociodemographic characteristics of the patients were assessed, their Psoriasis Area Severity Index (PASI) scores were calculated to determine the clinical severity of the psoriasis, and the values were recorded. In addition, Dermatology Life Quality Index (DLQI), Body Image Scale, and Rosenberg Self-Esteem Scale results were evaluated. When the control and psoriasis groups were evaluated regarding the DLQI, self-esteem, and body image, quality of life was found to be more negatively affected in the psoriasis group than the controls, which was statistically significant ( P < 0.001), and self-esteem ( P < 0.001) and body image ( P < 0.001) were found to be significantly lower. Educational status significantly affected self-esteem ( P < 0.001) and body image ( P = 0.021), however, quality of life was not significantly affected by this parameter ( P = 0.345). PASI was positively correlated with the quality of life ( r = 0.703) and self-esteem ( r = 0.448), however, it was negatively correlated with the body image ( r = -0.423). Psoriasis may negatively affect quality of life, self-esteem, and body image, and may also cause psychosocial problems. An assessment of new approaches on this issue may contribute to developments in the treatment of and rehabilitation from this disease.

  1. Characteristics of a New X-Ray Imaging System for Interventional Procedures: Improved Image Quality and Reduced Radiation Dose.

    PubMed

    Schernthaner, Ruediger E; Haroun, Reham R; Nguyen, Sonny; Duran, Rafael; Sohn, Jae Ho; Sahu, Sonia; Chapiro, Julius; Zhao, Yan; Radaelli, Alessandro; van der Bom, Imramsjah M; Mauti, Maria; Hong, Kelvin; Geschwind, Jean-François H; Lin, MingDe

    2018-03-01

    To compare image quality and radiation exposure between a new angiographic imaging system and the preceding generation system during uterine artery embolization (UAE). In this retrospective, IRB-approved two-arm study, 54 patients with symptomatic uterine fibroids were treated with UAE on two different angiographic imaging systems. The new system includes optimized acquisition parameters and real-time image processing algorithms. Air kerma (AK), dose area product (DAP) and acquisition time for digital fluoroscopy (DF) and digital subtraction angiography (DSA) were recorded. Body mass index was noted as well. DF image quality was assessed objectively by image noise measurements. DSA image quality was rated by two blinded, independent readers on a four-rank scale. Statistical differences were assessed with unpaired t tests and Wilcoxon rank-sum tests. There was no significant difference between the patients treated on the new (n = 36) and the old system (n = 18) regarding age (p = 0.10), BMI (p = 0.18), DF time (p = 0.35) and DSA time (p = 0.17). The new system significantly reduced the cumulative AK and DAP by 64 and 72%, respectively (median 0.58 Gy and 145.9 Gy*cm 2 vs. 1.62 Gy and 526.8 Gy*cm 2 , p < 0.01 for both). Specifically, DAP for DF and DSA decreased by 59% (75.3 vs. 181.9 Gy*cm 2 , p < 0.01) and 78% (67.6 vs. 312.2 Gy*cm 2 , p < 0.01), respectively. The new system achieved a significant decrease in DF image noise (p < 0.01) and a significantly better DSA image quality (p < 0.01). The new angiographic imaging system significantly improved image quality and reduced radiation exposure during UAE procedures.

  2. Evaluation of image quality of digital photo documentation of female genital injuries following sexual assault.

    PubMed

    Ernst, E J; Speck, Patricia M; Fitzpatrick, Joyce J

    2011-12-01

    With the patient's consent, physical injuries sustained in a sexual assault are evaluated and treated by the sexual assault nurse examiner (SANE) and documented on preprinted traumagrams and with photographs. Digital imaging is now available to the SANE for documentation of sexual assault injuries, but studies of the image quality of forensic digital imaging of female genital injuries after sexual assault were not found in the literature. The Photo Documentation Image Quality Scoring System (PDIQSS) was developed to rate the image quality of digital photo documentation of female genital injuries after sexual assault. Three expert observers performed evaluations on 30 separate images at two points in time. An image quality score, the sum of eight integral technical and anatomical attributes on the PDIQSS, was obtained for each image. Individual image quality ratings, defined by rating image quality for each of the data, were also determined. The results demonstrated a high level of image quality and agreement when measured in all dimensions. For the SANE in clinical practice, the results of this study indicate that a high degree of agreement exists between expert observers when using the PDIQSS to rate image quality of individual digital photographs of female genital injuries after sexual assault. © 2011 International Association of Forensic Nurses.

  3. Image quality stability of whole-body diffusion weighted imaging.

    PubMed

    Chen, Yun-bin; Hu, Chun-miao; Zhong, Jing; Sun, Fei

    2009-06-01

    To assess the reproducibility of whole-body diffusion weighted imaging (WB-DWI) technique in healthy volunteers under normal breathing with background body signal suppression. WB-DWI was performed on 32 healthy volunteers twice within two-week period using short TI inversion-recovery diffusion-weighted echo-planar imaging sequence and built-in body coil. The volunteers were scanned across six stations continuously covering the entire body from the head to the feet under normal breathing. The bone apparent diffusion coefficient (ADC) and exponential ADC (eADC) of regions of interest (ROIs) were measured. We analyzed correlation of the results using paired-t-test to assess the reproducibility of the WB-DWI technique. We were successful in collecting and analyzing data of 64 WB-DWI images. There was no significant difference in bone ADC and eADC of 824 ROIs between the paired observers and paired scans (P>0.05). Most of the images from all stations were of diagnostic quality. The measurements of bone ADC and eADC have good reproducibility. WB-DWI technique under normal breathing with background body signal suppression is adequate.

  4. The influence of software filtering in digital mammography image quality

    NASA Astrophysics Data System (ADS)

    Michail, C.; Spyropoulou, V.; Kalyvas, N.; Valais, I.; Dimitropoulos, N.; Fountos, G.; Kandarakis, I.; Panayiotakis, G.

    2009-05-01

    Breast cancer is one of the most frequently diagnosed cancers among women. Several techniques have been developed to help in the early detection of breast cancer such as conventional and digital x-ray mammography, positron and single-photon emission mammography, etc. A key advantage in digital mammography is that images can be manipulated as simple computer image files. Thus non-dedicated commercially available image manipulation software can be employed to process and store the images. The image processing tools of the Photoshop (CS 2) software usually incorporate digital filters which may be used to reduce image noise, enhance contrast and increase spatial resolution. However, improving an image quality parameter may result in degradation of another. The aim of this work was to investigate the influence of three sharpening filters, named hereafter sharpen, sharpen more and sharpen edges on image resolution and noise. Image resolution was assessed by means of the Modulation Transfer Function (MTF).In conclusion it was found that the correct use of commercial non-dedicated software on digital mammograms may improve some aspects of image quality.

  5. Bayesian framework inspired no-reference region-of-interest quality measure for brain MRI images

    PubMed Central

    Osadebey, Michael; Pedersen, Marius; Arnold, Douglas; Wendel-Mitoraj, Katrina

    2017-01-01

    Abstract. We describe a postacquisition, attribute-based quality assessment method for brain magnetic resonance imaging (MRI) images. It is based on the application of Bayes theory to the relationship between entropy and image quality attributes. The entropy feature image of a slice is segmented into low- and high-entropy regions. For each entropy region, there are three separate observations of contrast, standard deviation, and sharpness quality attributes. A quality index for a quality attribute is the posterior probability of an entropy region given any corresponding region in a feature image where quality attribute is observed. Prior belief in each entropy region is determined from normalized total clique potential (TCP) energy of the slice. For TCP below the predefined threshold, the prior probability for a region is determined by deviation of its percentage composition in the slice from a standard normal distribution built from 250 MRI volume data provided by Alzheimer’s Disease Neuroimaging Initiative. For TCP above the threshold, the prior is computed using a mathematical model that describes the TCP–noise level relationship in brain MRI images. Our proposed method assesses the image quality of each entropy region and the global image. Experimental results demonstrate good correlation with subjective opinions of radiologists for different types and levels of quality distortions. PMID:28630885

  6. “Lucky Averaging”: Quality improvement on Adaptive Optics Scanning Laser Ophthalmoscope Images

    PubMed Central

    Huang, Gang; Zhong, Zhangyi; Zou, Weiyao; Burns, Stephen A.

    2012-01-01

    Adaptive optics(AO) has greatly improved retinal image resolution. However, even with AO, temporal and spatial variations in image quality still occur due to wavefront fluctuations, intra-frame focus shifts and other factors. As a result, aligning and averaging images can produce a mean image that has lower resolution or contrast than the best images within a sequence. To address this, we propose an image post-processing scheme called “lucky averaging”, analogous to lucky imaging (Fried, 1978) based on computing the best local contrast over time. Results from eye data demonstrate improvements in image quality. PMID:21964097

  7. 5 CFR 850.301 - Electronic records; other acceptable records.

    Code of Federal Regulations, 2014 CFR

    2014-01-01

    ... SERVICE REGULATIONS (CONTINUED) ELECTRONIC RETIREMENT PROCESSING Records § 850.301 Electronic records; other acceptable records. (a) Acceptable electronic records for retirement and insurance processing by... (SF 2806 or SF 3100), or data or images obtained from such documents, including images stored in EDMS...

  8. An inter-laboratory comparison study of image quality of PET scanners using the NEMA NU 2-2001 procedure for assessment of image quality

    NASA Astrophysics Data System (ADS)

    Bergmann, Helmar; Dobrozemsky, Georg; Minear, Gregory; Nicoletti, Rudolf; Samal, Martin

    2005-05-01

    An inter-laboratory comparison study was conducted to assess the image quality of PET scanners in Austria. The survey included both dedicated PET scanners (D-PET, n = 8) and coincidence cameras (GC-PET, n = 7). Measurement of image quality was based on the NEMA (National Electrical Manufacturers Association) NU 2-2001 protocol and the IEC (International Electrotechnical Commission) body phantom. The latter contains six fillable spheres ranging in diameter from 37 mm down to 10 mm and a 'lung' insert. The two largest lesions L1-2 simulate cold lesions, the four smaller ones (L3-6) are filled with 18F and activity concentration ratios relative to background of 8:1 and 4:1, respectively. Acquisition and reconstruction in the study employed the participating institutes' standard oncological processing protocol. Calculation of contrast of the spheres was performed with a fully automated procedure. Contrast quality indices (CQIs) reflecting global performance were obtained by summing individual contrast values. Other image quality parameters calculated according to the NEMA protocol were background variability and relative error for correction of attenuation and scatter. Contrast values obtained were 61 ± 16 and 37 ± 14 for L1 (per cent contrast ± SD for D-PET and GC-PET, respectively), 57 ± 16 and 29 ± 16 for L2, 46 ± 10 and 26 ± 6.3 for L3, 37 ± 10 and 15 ± 4.3 for L4, 26 ± 11.5 and 6.1 ± 2.5 for L5, 14 ± 7.1 and 2.6 ± 2.6 for L6, with D-PET systems consistently being superior to GC-PET systems. CQIs permitted ranking of the scanners, also demonstrating a clear distinction between D-PET and GC-PET systems. Background variability was largest for GC-PET systems; the relative error of attenuation and scatter correction was significantly correlated with image quality for D-PET systems only. The study demonstrated considerable differences in image quality not only between GC-PET and D-PET systems but also between individual D-PET systems with possible

  9. [Psychological impact of ostomy on the quality of life of colorectal cancer patients: Role of body image, self-esteem and anxiety].

    PubMed

    Beaubrun En Famille Diant, Laury; Sordes, Florence; Chaubard, Thierry

    2018-06-01

    Surgery accounts for an important part of the therapeutic arsenal of colorectal cancer treatment. In digestive cancers, ostomy devices induce the loss of anal function and control. This medical appliance generates changes affecting all aspects of patients' lives. This study explores, on the one hand, the psychological impact of colostomy on colorectal cancer patients' quality of life and on the other hand, it analyzes the correlational links between body image, self-esteem and anxiety during the stoma. Thirty-five patients with colorectal cancer participated in the study, divided into 2 subgroups: 23 were carriers of a definitive stoma and the 12 others with a temporary stoma. All completed the Functional Assessment Cancer Therapy (FACT-C), the Body Image Scale (BIS), the State Trait Anxiety Inventory (STAI-Y) and the Self Esteem Scale (ETES). Analysis revealed the quality of life of temporary ostomates is more affected than that of the definitive ones. All three of them, body image, self-esteem and anxiety negatively affect the quality of life regardless of the type of stoma. This study highlights the prevalence of physical self-esteem for temporary ostomy; the role of a good body image and substantial emotional self-esteem for the permanent ostomy. Future studies are required to explore the underlying causes of the acceptance of this equipment and the mediating role of care devices. Copyright © 2018 Société Française du Cancer. Published by Elsevier Masson SAS. All rights reserved.

  10. Relationships among muscle dysmorphia characteristics, body image quality of life, and coping in males.

    PubMed

    Tod, D; Edwards, C

    2015-09-01

    The purpose of this study was to examine relationships among bodybuilding dependence, muscle satisfaction, body image-related quality of life and body image-related coping strategies, and test the hypothesis that muscle dysmorphia characteristics may predict quality of life via coping strategies. Participants (294 males, Mage=20.5 years, SD=3.1) participated in a cross-sectional survey. Participants completed questionnaires assessing muscle satisfaction, bodybuilding dependence, body image-related quality of life and body image-related coping. Quality of life was correlated positively with muscle satisfaction and bodybuilding dependence but negatively with body image coping (P<0.05). Body image coping was correlated positively with bodybuilding dependence and negatively with muscle satisfaction (P<0.05). Mediation analysis found that bodybuilding dependence and muscle satisfaction predicted quality of life both directly and indirectly via body image coping strategies (as evidenced by the bias corrected and accelerated bootstrapped confidence intervals). These results provide preliminary evidence regarding the ways that muscularity concerns might influence body image-related quality of life. Copyright © 2014 Sports Medicine Australia. Published by Elsevier Ltd. All rights reserved.

  11. Simultaneous fluoroscopic and nuclear imaging: impact of collimator choice on nuclear image quality.

    PubMed

    van der Velden, Sandra; Beijst, Casper; Viergever, Max A; de Jong, Hugo W A M

    2017-01-01

    X-ray-guided oncological interventions could benefit from the availability of simultaneously acquired nuclear images during the procedure. To this end, a real-time, hybrid fluoroscopic and nuclear imaging device, consisting of an X-ray c-arm combined with gamma imaging capability, is currently being developed (Beijst C, Elschot M, Viergever MA, de Jong HW. Radiol. 2015;278:232-238). The setup comprises four gamma cameras placed adjacent to the X-ray tube. The four camera views are used to reconstruct an intermediate three-dimensional image, which is subsequently converted to a virtual nuclear projection image that overlaps with the X-ray image. The purpose of the present simulation study is to evaluate the impact of gamma camera collimator choice (parallel hole versus pinhole) on the quality of the virtual nuclear image. Simulation studies were performed with a digital image quality phantom including realistic noise and resolution effects, with a dynamic frame acquisition time of 1 s and a total activity of 150 MBq. Projections were simulated for 3, 5, and 7 mm pinholes and for three parallel hole collimators (low-energy all-purpose (LEAP), low-energy high-resolution (LEHR) and low-energy ultra-high-resolution (LEUHR)). Intermediate reconstruction was performed with maximum likelihood expectation-maximization (MLEM) with point spread function (PSF) modeling. In the virtual projection derived therefrom, contrast, noise level, and detectability were determined and compared with the ideal projection, that is, as if a gamma camera were located at the position of the X-ray detector. Furthermore, image deformations and spatial resolution were quantified. Additionally, simultaneous fluoroscopic and nuclear images of a sphere phantom were acquired with a physical prototype system and compared with the simulations. For small hot spots, contrast is comparable for all simulated collimators. Noise levels are, however, 3 to 8 times higher in pinhole geometries than in parallel

  12. A cross-platform survey of CT image quality and dose from routine abdomen protocols and a method to systematically standardize image quality

    PubMed Central

    Favazza, Christopher P.; Duan, Xinhui; Zhang, Yi; Yu, Lifeng; Leng, Shuai; Kofler, James M.; Bruesewitz, Michael R.; McCollough, Cynthia H.

    2015-01-01

    Through this investigation we developed a methodology to evaluate and standardize CT image quality from routine abdomen protocols across different manufacturers and models. The influence of manufacturer-specific automated exposure control systems on image quality was directly assessed to standardize performance across a range of patient sizes. We evaluated 16 CT scanners across our health system, including Siemens, GE, and Toshiba models. Using each practice’s routine abdomen protocol, we measured spatial resolution, image noise, and scanner radiation output (CTDIvol). Axial and in-plane spatial resolutions were assessed through slice sensitivity profile (SSP) and modulation transfer function (MTF) measurements, respectively. Image noise and CTDIvol values were obtained for three different phantom sizes. SSP measurements demonstrated a bimodal distribution in slice widths: an average of 6.2 ± 0.2 mm using GE’s “Plus” mode reconstruction setting and 5.0 ± 0.1 mm for all other scanners. MTF curves were similar for all scanners. Average spatial frequencies at 50%, 10%, and 2% MTF values were 3.24 ± 0.37, 6.20 ± 0.34, and 7.84 ± 0.70 lp/cm, respectively. For all phantom sizes, image noise and CTDIvol varied considerably: 6.5–13.3 HU (noise) and 4.8–13.3 mGy (CTDIvol) for the smallest phantom; 9.1–18.4 HU and 9.3–28.8 mGy for the medium phantom; and 7.8–23.4 HU and 16.0–48.1 mGy for the largest phantom. Using these measurements and benchmark SSP, MTF, and image noise targets, CT image quality can be standardized across a range of patient sizes. PMID:26459751

  13. 48 CFR 2446.502 - Responsibility for acceptance.

    Code of Federal Regulations, 2010 CFR

    2010-10-01

    ... 48 Federal Acquisition Regulations System 6 2010-10-01 2010-10-01 true Responsibility for acceptance. 2446.502 Section 2446.502 Federal Acquisition Regulations System DEPARTMENT OF HOUSING AND URBAN DEVELOPMENT CONTRACT MANAGEMENT QUALITY ASSURANCE Acceptance 2446.502 Responsibility for acceptance. ...

  14. SU-E-J-87: Building Deformation Error Histogram and Quality Assurance of Deformable Image Registration.

    PubMed

    Park, S B; Kim, H; Yao, M; Ellis, R; Machtay, M; Sohn, J W

    2012-06-01

    To quantify the systematic error of a Deformable Image Registration (DIR) system and establish Quality Assurance (QA) procedure. To address the shortfall of landmark approach which it is only available at the significant visible feature points, we adapted a Deformation Vector Map (DVM) comparison approach. We used two CT image sets (R and T image sets) taken for the same patient at different time and generated a DVM, which includes the DIR systematic error. The DVM was calculated using fine-tuned B-Spline DIR and L-BFGS optimizer. By utilizing this DVM we generated R' image set to eliminate the systematic error in DVM,. Thus, we have truth data set, R' and T image sets, and the truth DVM. To test a DIR system, we use R' and T image sets to a DIR system. We compare the test DVM to the truth DVM. If there is no systematic error, they should be identical. We built Deformation Error Histogram (DEH) for quantitative analysis. The test registration was performed with an in-house B-Spline DIR system using a stochastic gradient descent optimizer. Our example data set was generated with a head and neck patient case. We also tested CT to CBCT deformable registration. We found skin regions which interface with the air has relatively larger errors. Also mobile joints such as shoulders had larger errors. Average error for ROIs were as follows; CTV: 0.4mm, Brain stem: 1.4mm, Shoulders: 1.6mm, and Normal tissues: 0.7mm. We succeeded to build DEH approach to quantify the DVM uncertainty. Our data sets are available for testing other systems in our web page. Utilizing DEH, users can decide how much systematic error they would accept. DEH and our data can be a tool for an AAPM task group to compose a DIR system QA guideline. This project is partially supported by the Agency for Healthcare Research and Quality (AHRQ) grant 1R18HS017424-01A2. © 2012 American Association of Physicists in Medicine.

  15. Phantom dosimetry and image quality of i-CAT FLX cone-beam computed tomography

    PubMed Central

    Ludlow, John B.; Walker, Cameron

    2013-01-01

    Introduction Increasing use of cone-beam computed tomography in orthodontics has been coupled with heightened concern with the long-term risks of x-ray exposure in orthodontic populations. An industry response to this has been to offer low-exposure alternative scanning options in newer cone-beam computed tomography models. Methods Effective doses resulting from various combinations of field size, and field location comparing child and adult anthropomorphic phantoms using the recently introduced i-CAT FLX cone-beam computed tomography unit were measured with Optical Stimulated Dosimetry using previously validated protocols. Scan protocols included High Resolution (360° rotation, 600 image frames, 120 kVp, 5 mA, 7.4 sec), Standard (360°, 300 frames, 120 kVp, 5 mA, 3.7 sec), QuickScan (180°, 160 frames, 120 kVp, 5 mA, 2 sec) and QuickScan+ (180°, 160 frames, 90 kVp, 3 mA, 2 sec). Contrast-to-noise ratio (CNR) was calculated as a quantitative measure of image quality for the various exposure options using the QUART DVT phantom. Results Child phantom doses were on average 36% greater than Adult phantom doses. QuickScan+ protocols resulted in significantly lower doses than Standard protocols for child (p=0.0167) and adult (p=0.0055) phantoms. 13×16 cm cephalometric fields of view ranged from 11–85 μSv in the adult phantom and 18–120 μSv in the child for QuickScan+ and Standard protocols respectively. CNR was reduced by approximately 2/3rds comparing QuickScan+ to Standard exposure parameters. Conclusions QuickScan+ effective doses are comparable to conventional panoramic examinations. Significant dose reductions are accompanied by significant reductions in image quality. However, this trade-off may be acceptable for certain diagnostic tasks such as interim assessment of treatment results. PMID:24286904

  16. Implementation of dictionary pair learning algorithm for image quality improvement

    NASA Astrophysics Data System (ADS)

    Vimala, C.; Aruna Priya, P.

    2018-04-01

    This paper proposes an image denoising on dictionary pair learning algorithm. Visual information is transmitted in the form of digital images is becoming a major method of communication in the modern age, but the image obtained after transmissions is often corrupted with noise. The received image needs processing before it can be used in applications. Image denoising involves the manipulation of the image data to produce a visually high quality image.

  17. Spread spectrum image watermarking based on perceptual quality metric.

    PubMed

    Zhang, Fan; Liu, Wenyu; Lin, Weisi; Ngan, King Ngi

    2011-11-01

    Efficient image watermarking calls for full exploitation of the perceptual distortion constraint. Second-order statistics of visual stimuli are regarded as critical features for perception. This paper proposes a second-order statistics (SOS)-based image quality metric, which considers the texture masking effect and the contrast sensitivity in Karhunen-Loève transform domain. Compared with the state-of-the-art metrics, the quality prediction by SOS better correlates with several subjectively rated image databases, in which the images are impaired by the typical coding and watermarking artifacts. With the explicit metric definition, spread spectrum watermarking is posed as an optimization problem: we search for a watermark to minimize the distortion of the watermarked image and to maximize the correlation between the watermark pattern and the spread spectrum carrier. The simple metric guarantees the optimal watermark a closed-form solution and a fast implementation. The experiments show that the proposed watermarking scheme can take full advantage of the distortion constraint and improve the robustness in return.

  18. Assessment of noise reduction potential and image quality improvement of a new generation adaptive statistical iterative reconstruction (ASIR-V) in chest CT.

    PubMed

    Tang, Hui; Yu, Nan; Jia, Yongjun; Yu, Yong; Duan, Haifeng; Han, Dong; Ma, Guangming; Ren, Chenglong; He, Taiping

    2018-01-01

    To evaluate the image quality improvement and noise reduction in routine dose, non-enhanced chest CT imaging by using a new generation adaptive statistical iterative reconstruction (ASIR-V) in comparison with ASIR algorithm. 30 patients who underwent routine dose, non-enhanced chest CT using GE Discovery CT750HU (GE Healthcare, Waukesha, WI) were included. The scan parameters included tube voltage of 120 kVp, automatic tube current modulation to obtain a noise index of 14HU, rotation speed of 0.6 s, pitch of 1.375:1 and slice thickness of 5 mm. After scanning, all scans were reconstructed with the recommended level of 40%ASIR for comparison purpose and different percentages of ASIR-V from 10% to 100% in a 10% increment. The CT attenuation values and SD of the subcutaneous fat, back muscle and descending aorta were measured at the level of tracheal carina of all reconstructed images. The signal-to-noise ratio (SNR) was calculated with SD representing image noise. The subjective image quality was independently evaluated by two experienced radiologists. For all ASIR-V images, the objective image noise (SD) of fat, muscle and aorta decreased and SNR increased along with increasing ASIR-V percentage. The SD of 30% ASIR-V to 100% ASIR-V was significantly lower than that of 40% ASIR (p < 0.05). In terms of subjective image evaluation, all ASIR-V reconstructions had good diagnostic acceptability. However, the 50% ASIR-V to 70% ASIR-V series showed significantly superior visibility of small structures when compared with the 40% ASIR and ASIR-V of other percentages (p < 0.05), and 60% ASIR-V was the best series of all ASIR-V images, with a highest subjective image quality. The image sharpness was significantly decreased in images reconstructed by 80% ASIR-V and higher. In routine dose, non-enhanced chest CT, ASIR-V shows greater potential in reducing image noise and artefacts and maintaining image sharpness when compared to the recommended level of 40%ASIR algorithm

  19. Improving Secondary Ion Mass Spectrometry Image Quality with Image Fusion

    PubMed Central

    Tarolli, Jay G.; Jackson, Lauren M.; Winograd, Nicholas

    2014-01-01

    The spatial resolution of chemical images acquired with cluster secondary ion mass spectrometry (SIMS) is limited not only by the size of the probe utilized to create the images, but also by detection sensitivity. As the probe size is reduced to below 1 µm, for example, a low signal in each pixel limits lateral resolution due to counting statistics considerations. Although it can be useful to implement numerical methods to mitigate this problem, here we investigate the use of image fusion to combine information from scanning electron microscope (SEM) data with chemically resolved SIMS images. The advantage of this approach is that the higher intensity and, hence, spatial resolution of the electron images can help to improve the quality of the SIMS images without sacrificing chemical specificity. Using a pan-sharpening algorithm, the method is illustrated using synthetic data, experimental data acquired from a metallic grid sample, and experimental data acquired from a lawn of algae cells. The results show that up to an order of magnitude increase in spatial resolution is possible to achieve. A cross-correlation metric is utilized for evaluating the reliability of the procedure. PMID:24912432

  20. Improving Secondary Ion Mass Spectrometry Image Quality with Image Fusion

    NASA Astrophysics Data System (ADS)

    Tarolli, Jay G.; Jackson, Lauren M.; Winograd, Nicholas

    2014-12-01

    The spatial resolution of chemical images acquired with cluster secondary ion mass spectrometry (SIMS) is limited not only by the size of the probe utilized to create the images but also by detection sensitivity. As the probe size is reduced to below 1 μm, for example, a low signal in each pixel limits lateral resolution because of counting statistics considerations. Although it can be useful to implement numerical methods to mitigate this problem, here we investigate the use of image fusion to combine information from scanning electron microscope (SEM) data with chemically resolved SIMS images. The advantage of this approach is that the higher intensity and, hence, spatial resolution of the electron images can help to improve the quality of the SIMS images without sacrificing chemical specificity. Using a pan-sharpening algorithm, the method is illustrated using synthetic data, experimental data acquired from a metallic grid sample, and experimental data acquired from a lawn of algae cells. The results show that up to an order of magnitude increase in spatial resolution is possible to achieve. A cross-correlation metric is utilized for evaluating the reliability of the procedure.

  1. 40 CFR 86.610-98 - Compliance with acceptable quality level and passing and failing criteria for Selective...

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... level and passing and failing criteria for Selective Enforcement Audits. 86.610-98 Section 86.610-98... and passing and failing criteria for Selective Enforcement Audits. (a) The prescribed acceptable quality level is 40 percent. (b) A failed vehicle is one whose final deteriorated test results pursuant to...

  2. MRIQC: Advancing the automatic prediction of image quality in MRI from unseen sites

    PubMed Central

    2017-01-01

    Quality control of MRI is essential for excluding problematic acquisitions and avoiding bias in subsequent image processing and analysis. Visual inspection is subjective and impractical for large scale datasets. Although automated quality assessments have been demonstrated on single-site datasets, it is unclear that solutions can generalize to unseen data acquired at new sites. Here, we introduce the MRI Quality Control tool (MRIQC), a tool for extracting quality measures and fitting a binary (accept/exclude) classifier. Our tool can be run both locally and as a free online service via the OpenNeuro.org portal. The classifier is trained on a publicly available, multi-site dataset (17 sites, N = 1102). We perform model selection evaluating different normalization and feature exclusion approaches aimed at maximizing across-site generalization and estimate an accuracy of 76%±13% on new sites, using leave-one-site-out cross-validation. We confirm that result on a held-out dataset (2 sites, N = 265) also obtaining a 76% accuracy. Even though the performance of the trained classifier is statistically above chance, we show that it is susceptible to site effects and unable to account for artifacts specific to new sites. MRIQC performs with high accuracy in intra-site prediction, but performance on unseen sites leaves space for improvement which might require more labeled data and new approaches to the between-site variability. Overcoming these limitations is crucial for a more objective quality assessment of neuroimaging data, and to enable the analysis of extremely large and multi-site samples. PMID:28945803

  3. No-reference image quality assessment based on statistics of convolution feature maps

    NASA Astrophysics Data System (ADS)

    Lv, Xiaoxin; Qin, Min; Chen, Xiaohui; Wei, Guo

    2018-04-01

    We propose a Convolutional Feature Maps (CFM) driven approach to accurately predict image quality. Our motivation bases on the finding that the Nature Scene Statistic (NSS) features on convolution feature maps are significantly sensitive to distortion degree of an image. In our method, a Convolutional Neural Network (CNN) is trained to obtain kernels for generating CFM. We design a forward NSS layer which performs on CFM to better extract NSS features. The quality aware features derived from the output of NSS layer is effective to describe the distortion type and degree an image suffered. Finally, a Support Vector Regression (SVR) is employed in our No-Reference Image Quality Assessment (NR-IQA) model to predict a subjective quality score of a distorted image. Experiments conducted on two public databases demonstrate the promising performance of the proposed method is competitive to state of the art NR-IQA methods.

  4. JPEG vs. JPEG 2000: an objective comparison of image encoding quality

    NASA Astrophysics Data System (ADS)

    Ebrahimi, Farzad; Chamik, Matthieu; Winkler, Stefan

    2004-11-01

    This paper describes an objective comparison of the image quality of different encoders. Our approach is based on estimating the visual impact of compression artifacts on perceived quality. We present a tool that measures these artifacts in an image and uses them to compute a prediction of the Mean Opinion Score (MOS) obtained in subjective experiments. We show that the MOS predictions by our proposed tool are a better indicator of perceived image quality than PSNR, especially for highly compressed images. For the encoder comparison, we compress a set of 29 test images with two JPEG encoders (Adobe Photoshop and IrfanView) and three JPEG2000 encoders (JasPer, Kakadu, and IrfanView) at various compression ratios. We compute blockiness, blur, and MOS predictions as well as PSNR of the compressed images. Our results show that the IrfanView JPEG encoder produces consistently better images than the Adobe Photoshop JPEG encoder at the same data rate. The differences between the JPEG2000 encoders in our test are less pronounced; JasPer comes out as the best codec, closely followed by IrfanView and Kakadu. Comparing the JPEG- and JPEG2000-encoding quality of IrfanView, we find that JPEG has a slight edge at low compression ratios, while JPEG2000 is the clear winner at medium and high compression ratios.

  5. Safety and Patient Acceptability of Stellate Ganglion Blockade as a Treatment Adjunct for Combat-Related Post-Traumatic Stress Disorder: A Quality Assurance Initiative.

    PubMed

    McLean, Brian

    2015-09-10

     To perform a quality assurance and performance improvement project through review of our single center data on the safety and patient acceptability of the stellate ganglion blockade (SGB) procedure for the relief of symptoms related to chronic post-traumatic stress disorder.  Our interventional pain management service has been offering trials of SGB therapy to assist with the management of the sympathetically mediated anxiety and hyperarousal symptoms of severe and treatment-refractory combat-related PTSD. There have been multiple case series in the literature describing the potential impact of this procedure for PTSD symptom management as well as the safety of image-guided procedures. We wished to ensure that we were performing this procedure safely and that patients were tolerating and accepting of this adjunctive treatment option.  We conducted a review of our quality assurance and performance improvement data over the past 18 months during which we performed 250 stellate ganglion blocks for the management of PTSD symptoms to detect any potential complications or unanticipated side effects.  We also analyzed responses from an anonymous patient de-identified survey collected regarding the comfort and satisfaction associated with the procedure.  We did not identify any immediate post-procedural complications or delayed complications from any of the 250 procedures performed from November 2013 to April 2015. Of the 110 surveys that were returned and tabulated, 100% of the patients surveyed were overall satisfied with our process and with the procedure, 100% said they would recommend the procedure to a friend, and 95% stated that they would be willing to undergo as many repeat procedures as necessary based on little discomfort and tolerable side effects.  Our quality assurance assessment suggests that in our center the SGB procedure for PTSD is a safe, well-tolerated, and acceptable treatment adjunct in the management of

  6. Image quality evaluation of eight complementary metal-oxide semiconductor intraoral digital X-ray sensors.

    PubMed

    Teich, Sorin; Al-Rawi, Wisam; Heima, Masahiro; Faddoul, Fady F; Goldzweig, Gil; Gutmacher, Zvi; Aizenbud, Dror

    2016-10-01

    To evaluate the image quality generated by eight commercially available intraoral sensors. Eighteen clinicians ranked the quality of a bitewing acquired from one subject using eight different intraoral sensors. Analytical methods used to evaluate clinical image quality included the Visual Grading Characteristics method, which helps to quantify subjective opinions to make them suitable for analysis. The Dexis sensor was ranked significantly better than Sirona and Carestream-Kodak sensors; and the image captured using the Carestream-Kodak sensor was ranked significantly worse than those captured using Dexis, Schick and Cyber Medical Imaging sensors. The Image Works sensor image was rated the lowest by all clinicians. Other comparisons resulted in non-significant results. None of the sensors was considered to generate images of significantly better quality than the other sensors tested. Further research should be directed towards determining the clinical significance of the differences in image quality reported in this study. © 2016 FDI World Dental Federation.

  7. Digital mammography--DQE versus optimized image quality in clinical environment: an on site study

    NASA Astrophysics Data System (ADS)

    Oberhofer, Nadia; Fracchetti, Alessandro; Springeth, Margareth; Moroder, Ehrenfried

    2010-04-01

    The intrinsic quality of the detection system of 7 different digital mammography units (5 direct radiography DR; 2 computed radiography CR), expressed by DQE, has been compared with their image quality/dose performances in clinical use. DQE measurements followed IEC 62220-1-2 using a tungsten test object for MTF determination. For image quality assessment two different methods have been applied: 1) measurement of contrast to noise ratio (CNR) according to the European guidelines and 2) contrast-detail (CD) evaluation. The latter was carried out with the phantom CDMAM ver. 3.4 and the commercial software CDMAM Analyser ver. 1.1 (both Artinis) for automated image analysis. The overall image quality index IQFinv proposed by the software has been validated. Correspondence between the two methods has been shown figuring out a linear correlation between CNR and IQFinv. All systems were optimized with respect to image quality and average glandular dose (AGD) within the constraints of automatic exposure control (AEC). For each equipment, a good image quality level was defined by means of CD analysis, and the corresponding CNR value considered as target value. The goal was to achieve for different PMMA-phantom thicknesses constant image quality, that means the CNR target value, at minimum dose. All DR systems exhibited higher DQE and significantly better image quality compared to CR systems. Generally switching, where available, to a target/filter combination with an x-ray spectrum of higher mean energy permitted dose savings at equal image quality. However, several systems did not allow to modify the AEC in order to apply optimal radiographic technique in clinical use. The best ratio image quality/dose was achieved by a unit with a-Se detector and W anode only recently available on the market.

  8. Paediatric cerebrovascular CT angiography-towards better image quality.

    PubMed

    Thust, Stefanie C; Chong, Wui Khean Kling; Gunny, Roxana; Mazumder, Asif; Poitelea, Marius; Welsh, Anna; Ederies, Ash; Mankad, Kshitij

    2014-12-01

    Paediatric cerebrovascular CT angiography (CTA) can be challenging to perform due to variable cardiovascular physiology between different age groups and the risk of movement artefact. This analysis aimed to determine what proportion of CTA at our institution was of diagnostic quality and identify technical factors which could be improved. a retrospective analysis of 20 cases was performed at a national paediatric neurovascular centre assessing image quality with a subjective scoring system and Hounsfield Unit (HU) measurements. Demographic data, contrast dose, flow rate and triggering times were recorded for each patient. Using a qualitative scoring system, 75% of studies were found to be of diagnostic quality (n=9 'good', n=6 'satisfactory') and 25% (n=5) were 'poor'. Those judged subjectively to be poor had arterial contrast density measured at less than 250 HU. Increased arterial opacification was achieved for cases performed with an increased flow rate (2.5-4 mL/s) and higher intravenous contrast dose (2 mL/kg). Triggering was found to be well timed in nine cases, early in four cases and late in seven cases. Of the scans triggered early, 75% were poor. Of the scans triggered late, less (29%) were poor. High flow rates (>2.5 mL/s) were a key factor for achieving high quality paediatric cerebrovascular CTA imaging. However, appropriate triggering by starting the scan immediately on contrast opacification of the monitoring vessel plays an important role and could maintain image quality when flow rates were lower. Early triggering appeared more detrimental than late.

  9. Image quality assessment for video stream recognition systems

    NASA Astrophysics Data System (ADS)

    Chernov, Timofey S.; Razumnuy, Nikita P.; Kozharinov, Alexander S.; Nikolaev, Dmitry P.; Arlazarov, Vladimir V.

    2018-04-01

    Recognition and machine vision systems have long been widely used in many disciplines to automate various processes of life and industry. Input images of optical recognition systems can be subjected to a large number of different distortions, especially in uncontrolled or natural shooting conditions, which leads to unpredictable results of recognition systems, making it impossible to assess their reliability. For this reason, it is necessary to perform quality control of the input data of recognition systems, which is facilitated by modern progress in the field of image quality evaluation. In this paper, we investigate the approach to designing optical recognition systems with built-in input image quality estimation modules and feedback, for which the necessary definitions are introduced and a model for describing such systems is constructed. The efficiency of this approach is illustrated by the example of solving the problem of selecting the best frames for recognition in a video stream for a system with limited resources. Experimental results are presented for the system for identity documents recognition, showing a significant increase in the accuracy and speed of the system under simulated conditions of automatic camera focusing, leading to blurring of frames.

  10. Improving lateral resolution and image quality of optical coherence tomography by the multi-frame superresolution technique for 3D tissue imaging

    PubMed Central

    Shen, Kai; Lu, Hui; Baig, Sarfaraz; Wang, Michael R.

    2017-01-01

    The multi-frame superresolution technique is introduced to significantly improve the lateral resolution and image quality of spectral domain optical coherence tomography (SD-OCT). Using several sets of low resolution C-scan 3D images with lateral sub-spot-spacing shifts on different sets, the multi-frame superresolution processing of these sets at each depth layer reconstructs a higher resolution and quality lateral image. Layer by layer processing yields an overall high lateral resolution and quality 3D image. In theory, the superresolution processing including deconvolution can solve the diffraction limit, lateral scan density and background noise problems together. In experiment, the improved lateral resolution by ~3 times reaching 7.81 µm and 2.19 µm using sample arm optics of 0.015 and 0.05 numerical aperture respectively as well as doubling the image quality has been confirmed by imaging a known resolution test target. Improved lateral resolution on in vitro skin C-scan images has been demonstrated. For in vivo 3D SD-OCT imaging of human skin, fingerprint and retina layer, we used the multi-modal volume registration method to effectively estimate the lateral image shifts among different C-scans due to random minor unintended live body motion. Further processing of these images generated high lateral resolution 3D images as well as high quality B-scan images of these in vivo tissues. PMID:29188089

  11. Improving lateral resolution and image quality of optical coherence tomography by the multi-frame superresolution technique for 3D tissue imaging.

    PubMed

    Shen, Kai; Lu, Hui; Baig, Sarfaraz; Wang, Michael R

    2017-11-01

    The multi-frame superresolution technique is introduced to significantly improve the lateral resolution and image quality of spectral domain optical coherence tomography (SD-OCT). Using several sets of low resolution C-scan 3D images with lateral sub-spot-spacing shifts on different sets, the multi-frame superresolution processing of these sets at each depth layer reconstructs a higher resolution and quality lateral image. Layer by layer processing yields an overall high lateral resolution and quality 3D image. In theory, the superresolution processing including deconvolution can solve the diffraction limit, lateral scan density and background noise problems together. In experiment, the improved lateral resolution by ~3 times reaching 7.81 µm and 2.19 µm using sample arm optics of 0.015 and 0.05 numerical aperture respectively as well as doubling the image quality has been confirmed by imaging a known resolution test target. Improved lateral resolution on in vitro skin C-scan images has been demonstrated. For in vivo 3D SD-OCT imaging of human skin, fingerprint and retina layer, we used the multi-modal volume registration method to effectively estimate the lateral image shifts among different C-scans due to random minor unintended live body motion. Further processing of these images generated high lateral resolution 3D images as well as high quality B-scan images of these in vivo tissues.

  12. Half-Fan-Based Intensity-Weighted Region-of-Interest Imaging for Low-Dose Cone-Beam CT in Image-Guided Radiation Therapy.

    PubMed

    Yoo, Boyeol; Son, Kihong; Pua, Rizza; Kim, Jinsung; Solodov, Alexander; Cho, Seungryong

    2016-10-01

    With the increased use of computed tomography (CT) in clinics, dose reduction is the most important feature people seek when considering new CT techniques or applications. We developed an intensity-weighted region-of-interest (IWROI) imaging method in an exact half-fan geometry to reduce the imaging radiation dose to patients in cone-beam CT (CBCT) for image-guided radiation therapy (IGRT). While dose reduction is highly desirable, preserving the high-quality images of the ROI is also important for target localization in IGRT. An intensity-weighting (IW) filter made of copper was mounted in place of a bowtie filter on the X-ray tube unit of an on-board imager (OBI) system such that the filter can substantially reduce radiation exposure to the outer ROI. In addition to mounting the IW filter, the lead-blade collimation of the OBI was adjusted to produce an exact half-fan scanning geometry for a further reduction of the radiation dose. The chord-based rebinned backprojection-filtration (BPF) algorithm in circular CBCT was implemented for image reconstruction, and a humanoid pelvis phantom was used for the IWROI imaging experiment. The IWROI image of the phantom was successfully reconstructed after beam-quality correction, and it was registered to the reference image within an acceptable level of tolerance. Dosimetric measurements revealed that the dose is reduced by approximately 61% in the inner ROI and by 73% in the outer ROI compared to the conventional bowtie filter-based half-fan scan. The IWROI method substantially reduces the imaging radiation dose and provides reconstructed images with an acceptable level of quality for patient setup and target localization. The proposed half-fan-based IWROI imaging technique can add a valuable option to CBCT in IGRT applications.

  13. SU-D-204-06: Dose and Image Quality Evaluation of a Low-Dose Slot-Scanning X-Ray System for Pediatric Orthopedic Studies

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Liu, Z; Hoerner, M; Lamoureux, R

    Purpose: Children in early teens with scoliosis require repeated radiographic exams over a number of years. The EOS (EOS imaging S.A., Paris, France) is a novel low-dose slot-scanning digital radiographic system designed to produce full-spine images of a free-standing patient. The radiation dose and image quality characteristics of the EOS were evaluated relative to those of a Computed Radiography (CR) system for scoliosis imaging. Methods: For dose evaluation, a full-torso anthropomorphic phantom was scanned five times using the default standard clinical protocols for both the EOS and a CR system, which include both posteroanterior and lateral full-spine views. Optically stimulatedmore » luminescent dosimeters (OSLDs), also known as nanoDots™ (Landauer, Inc., Glenwood, IL), were placed on the phantom’s surface to measure entrance skin dose. To assess image quality, MTF curves were generated from sampling the noise levels within the high-contrast regions of a line-pair phantom. Vertical and horizontal distortions were measured for the square line-pair phantom with the EOS system to evaluate the effects of geometric magnification and misalignment with the indicated imaging plane. Results: The entrance skin dose was measured to be 0.4 to 1.1 mGy for the EOS, and 0.7 to 3.6 mGy for the CR study. MTF comparison shows that CR greatly outperforms the EOS, despite both systems having a limiting resolution at 1.8 line-pairs per mm. Vertical distortion was unaffected by phantom positioning, because of the EOS slot-scanning geometry. Horizontal distortion increased linearly with miscentering distance. Conclusion: The EOS system resulted in approximately 70% lower radiation dose than CR for full-spine images. Image quality was found to be inferior to CR. Further investigation is required to see if EOS system is an acceptable modality for performing clinically diagnostic scoliosis examinations.« less

  14. Evaluation of the Spies™ modalities image quality

    PubMed Central

    Emiliani, Esteban; Talso, Michele; Baghdadi, Mohammed; Barreiro, Aarón; Orosa, Andrea; Serviàn, Pol; Gavrilov, Pavel; Proietti, Silvia; Traxer, Olivier

    2017-01-01

    Introduction The Spies™ system (Karl-Storz®) was introduced into digital ureteroscopy to improve endoscopic vision. To date, there is no data to either indicate which of the Spies modalities is better for improving diagnosis and treatment procedures, nor to compare the modalities in terms of image quality. The aim of this study was to evaluate and compare the image quality of five Spies™ modalities (SM) to the standard white light in an in-vitro model. Materials and Methods Two standardized grids and 3 stones of different composition were recorded in white light and the 5SM (Clara, Chroma, Clara+Chroma), Spectra A and B) using 4 standardized aqueous scenarios. Twelve templates were done in order to simultaneously compare the same objective in the different modalities. Six urologists, five medical students, five urology residents, and five persons not involved with urology evaluated each video on a scale of 1 (very bad) to 5 (very good). Results Comparing white light to SM, subjects scored better the quality of Clara and Clara+Chroma than white light (p=0.0139 and p<0.05) and scored worse Spectra A and B (p=0.0005 and p=0.0023)). When comparing Clara to the other SM, it was ranked equivalent to Clara+Chroma (p=0.67) and obtained a higher rank than Chroma, Spectra A and B (p<0.05, p=0.0001 and p=0.0001). In the multivariate analysis mean scores were higher among urologists. Conclusion In all analyzed scenarios, the subjects ranked Clara and Clara+Chroma as the modalities with better image quality compared to white light. PMID:28338307

  15. Sliding window adaptive histogram equalization of intraoral radiographs: effect on image quality.

    PubMed

    Sund, T; Møystad, A

    2006-05-01

    To investigate whether contrast enhancement by non-interactive, sliding window adaptive histogram equalization (SWAHE) can enhance the image quality of intraoral radiographs in the dental clinic. Three dentists read 22 periapical and 12 bitewing storage phosphor (SP) radiographs. For the periapical readings they graded the quality of the examination with regard to visually locating the root apex. For the bitewing readings they registered all occurrences of approximal caries on a confidence scale. Each reading was first done on an unprocessed radiograph ("single-view"), and then re-done with the image processed with SWAHE displayed beside the unprocessed version ("twin-view"). The processing parameters for SWAHE were the same for all the images. For the periapical examinations, twin-view was judged to raise the image quality for 52% of those cases where the single-view quality was below the maximum. For the bitewing radiographs, there was a change of caries classification (both positive and negative) with twin-view in 19% of the cases, but with only a 3% net increase in the total number of caries registrations. For both examinations interobserver variance was unaffected. Non-interactive SWAHE applied to dental SP radiographs produces a supplemental contrast enhanced image which in twin-view reading improves the image quality of periapical examinations. SWAHE also affects caries diagnosis of bitewing images, and further study using a gold standard is warranted.

  16. INCITS W1.1 development update: appearance-based image quality standards for printers

    NASA Astrophysics Data System (ADS)

    Zeise, Eric K.; Rasmussen, D. René; Ng, Yee S.; Dalal, Edul; McCarthy, Ann; Williams, Don

    2008-01-01

    In September 2000, INCITS W1 (the U.S. representative of ISO/IEC JTC1/SC28, the standardization committee for office equipment) was chartered to develop an appearance-based image quality standard. (1),(2) The resulting W1.1 project is based on a proposal (3) that perceived image quality can be described by a small set of broad-based attributes. There are currently six ad hoc teams, each working towards the development of standards for evaluation of perceptual image quality of color printers for one or more of these image quality attributes. This paper summarizes the work in progress of the teams addressing the attributes of Macro-Uniformity, Colour Rendition, Gloss & Gloss Uniformity, Text & Line Quality and Effective Resolution.

  17. Integrating image quality in 2nu-SVM biometric match score fusion.

    PubMed

    Vatsa, Mayank; Singh, Richa; Noore, Afzel

    2007-10-01

    This paper proposes an intelligent 2nu-support vector machine based match score fusion algorithm to improve the performance of face and iris recognition by integrating the quality of images. The proposed algorithm applies redundant discrete wavelet transform to evaluate the underlying linear and non-linear features present in the image. A composite quality score is computed to determine the extent of smoothness, sharpness, noise, and other pertinent features present in each subband of the image. The match score and the corresponding quality score of an image are fused using 2nu-support vector machine to improve the verification performance. The proposed algorithm is experimentally validated using the FERET face database and the CASIA iris database. The verification performance and statistical evaluation show that the proposed algorithm outperforms existing fusion algorithms.

  18. Impact of orthodontic appliances on the quality of craniofacial anatomical magnetic resonance imaging and real-time speech imaging.

    PubMed

    Wylezinska, Marzena; Pinkstone, Marie; Hay, Norman; Scott, Andrew D; Birch, Malcolm J; Miquel, Marc E

    2015-12-01

    The aim of this work was to investigate the effects of commonly used orthodontic appliances on the magnetic resonance (MR) image quality of the craniofacial region, with special interest in the soft palate and velopharyngeal wall using real-time speech imaging sequences and anatomical imaging of the temporomandibular joints (TMJ) and pituitaries. Common orthodontic appliances were studied on 1.5 T scanner using standard spin and gradient echo sequences (based on the American Society for Testing and Materials standard test method) and sequences previously applied for high-resolution anatomical and dynamic real-time imaging during speech. Images were evaluated for the presence and size of artefacts. Metallic orthodontic appliances had different effects on image quality. The most extensive individual effects were associated with the presence of stainless steel archwire, particularly if combined with stainless steel brackets and stainless steel molar bands. With those appliances, diagnostic quality of magnetic resonance imaging speech and palate images will be most likely severely degraded, or speech imaging and imaging of pituitaries and TMJ will be not possible. All non-metallic, non-metallic with Ni/Cr reinforcement or Ni/Ti alloys appliances were of little concern. The results in the study are only valid at 1.5 T and for the sequences and devices used and cannot necessarily be extrapolated to all sequences and devices. Furthermore, both geometry and size of some appliances are subject dependent, and consequently, the effects on the image quality can vary between subjects. Therefore, the results presented in this article should be treated as a guide when assessing the risks of image quality degradation rather than an absolute evaluation of possible artefacts. Appliances manufactured from stainless steel cause extensive artefacts, which may render image non-diagnostic. The presence and type of orthodontic appliances should be always included in the patient

  19. Effect of attenuation correction on image quality in emission tomography

    NASA Astrophysics Data System (ADS)

    Denisova, N. V.; Ondar, M. M.

    2017-10-01

    In this paper, mathematical modeling and computer simulations of myocardial perfusion SPECT imaging are performed. The main factors affecting the quality of reconstructed images in SPECT are anatomical structures, the diastolic volume of a myocardium and attenuation of gamma rays. The purpose of the present work is to study the effect of attenuation correction on image quality in emission tomography. The basic 2D model describing a Tc-99m distribution in a transaxial slice of the thoracic part of a patient body was designed. This model was used to construct four phantoms simulated various anatomical shapes: 2 male and 2 female patients with normal, obese and subtle physique were included in the study. Data acquisition model which includes the effect of non-uniform attenuation, collimator-detector response and Poisson statistics was developed. The projection data were calculated for 60 views in accordance with the standard myocardial perfusion SPECT imaging protocol. Reconstructions of images were performed using the OSEM algorithm which is widely used in modern SPECT systems. Two types of patient's examination procedures were simulated: SPECT without attenuation correction and SPECT/CT with attenuation correction. The obtained results indicate a significant effect of the attenuation correction on the SPECT images quality.

  20. An in vitro comparison of subjective image quality of panoramic views acquired via 2D or 3D imaging.

    PubMed

    Pittayapat, P; Galiti, D; Huang, Y; Dreesen, K; Schreurs, M; Souza, P Couto; Rubira-Bullen, I R F; Westphalen, F H; Pauwels, R; Kalema, G; Willems, G; Jacobs, R

    2013-01-01

    The objective of this study is to compare subjective image quality and diagnostic validity of cone-beam CT (CBCT) panoramic reformatting with digital panoramic radiographs. Four dry human skulls and two formalin-fixed human heads were scanned using nine different CBCTs, one multi-slice CT (MSCT) and one standard digital panoramic device. Panoramic views were generated from CBCTs in four slice thicknesses. Seven observers scored image quality and visibility of 14 anatomical structures. Four observers repeated the observation after 4 weeks. Digital panoramic radiographs showed significantly better visualization of anatomical structures except for the condyle. Statistical analysis of image quality showed that the 3D imaging modalities (CBCTs and MSCT) were 7.3 times more likely to receive poor scores than the 2D modality. Yet, image quality from NewTom VGi® and 3D Accuitomo 170® was almost equivalent to that of digital panoramic radiographs with respective odds ratio estimates of 1.2 and 1.6 at 95% Wald confidence limits. A substantial overall agreement amongst observers was found. Intra-observer agreement was moderate to substantial. While 2D-panoramic images are significantly better for subjective diagnosis, 2/3 of the 3D-reformatted panoramic images are moderate or good for diagnostic purposes. Panoramic reformattings from particular CBCTs are comparable to digital panoramic images concerning the overall image quality and visualization of anatomical structures. This clinically implies that a 3D-derived panoramic view can be generated for diagnosis with a recommended 20-mm slice thickness, if CBCT data is a priori available for other purposes.

  1. 48 CFR 1346.503 - Place of acceptance.

    Code of Federal Regulations, 2010 CFR

    2010-10-01

    ... 48 Federal Acquisition Regulations System 5 2010-10-01 2010-10-01 false Place of acceptance. 1346.503 Section 1346.503 Federal Acquisition Regulations System DEPARTMENT OF COMMERCE CONTRACT MANAGEMENT QUALITY ASSURANCE Acceptance 1346.503 Place of acceptance. Insert a clause substantially similar to 1352...

  2. Task-based measures of image quality and their relation to radiation dose and patient risk

    PubMed Central

    Barrett, Harrison H.; Myers, Kyle J.; Hoeschen, Christoph; Kupinski, Matthew A.; Little, Mark P.

    2015-01-01

    The theory of task-based assessment of image quality is reviewed in the context of imaging with ionizing radiation, and objective figures of merit (FOMs) for image quality are summarized. The variation of the FOMs with the task, the observer and especially with the mean number of photons recorded in the image is discussed. Then various standard methods for specifying radiation dose are reviewed and related to the mean number of photons in the image and hence to image quality. Current knowledge of the relation between local radiation dose and the risk of various adverse effects is summarized, and some graphical depictions of the tradeoffs between image quality and risk are introduced. Then various dose-reduction strategies are discussed in terms of their effect on task-based measures of image quality. PMID:25564960

  3. Healthy Body Image Intervention Delivered to Young Women via Facebook Groups: Formative Study of Engagement and Acceptability.

    PubMed

    Stapleton, Jerod L; Manne, Sharon L; Day, Ashley K; Levonyan-Radloff, Kristine; Pagoto, Sherry L

    2018-02-20

    There is increasing interest in using social media sites such as Facebook to deliver health interventions so as to expose people to content while they are engaging in their usual social media habit. This formative intervention development study is novel in describing a preliminary test of using the secret group feature of Facebook to deliver a behavioral intervention targeting users of indoor tanning beds to reduce their risk of skin cancer. Intervention content was designed to challenge body image-related constructs associated with indoor tanning through the use of dissonance-inducing content. To evaluate engagement with and acceptability of using a secret Facebook group to deliver a healthy body image intervention to young women engaged in indoor tanning. Seventeen young women completed a baseline survey and joined a secret Facebook group with intervention content delivered via daily posts for 4 weeks. Engagement data was extracted and acceptability was measured via a follow-up survey. The study had a high retention rate (94%, [16/17]). On average, posts were viewed by 91% of participants, liked by 35%, and commented on by 26%. The average comment rate was highest (65%) for posts that elicited comments by directly posing questions or discussion topics to the group. Average intervention acceptability ratings were highly positive and participants reported feeling connected to the group and its topic. Average rates of past 1-month indoor tanning reported following the intervention were lower than the baseline rate (P=.08, Cohen d=0.47). This study is novel in demonstrating participant engagement with and acceptability of using Facebook secret groups to deliver a dissonance-inducing intervention approach that utilizes group-based discussions related to body image. The study is also unique within the field of skin cancer prevention by demonstrating the potential value of delivering an indoor tanning intervention within an interactive social media format. The findings

  4. Healthy Body Image Intervention Delivered to Young Women via Facebook Groups: Formative Study of Engagement and Acceptability

    PubMed Central

    Manne, Sharon L; Day, Ashley K; Levonyan-Radloff, Kristine; Pagoto, Sherry L

    2018-01-01

    Background There is increasing interest in using social media sites such as Facebook to deliver health interventions so as to expose people to content while they are engaging in their usual social media habit. This formative intervention development study is novel in describing a preliminary test of using the secret group feature of Facebook to deliver a behavioral intervention targeting users of indoor tanning beds to reduce their risk of skin cancer. Intervention content was designed to challenge body image-related constructs associated with indoor tanning through the use of dissonance-inducing content. Objective To evaluate engagement with and acceptability of using a secret Facebook group to deliver a healthy body image intervention to young women engaged in indoor tanning. Methods Seventeen young women completed a baseline survey and joined a secret Facebook group with intervention content delivered via daily posts for 4 weeks. Engagement data was extracted and acceptability was measured via a follow-up survey. Results The study had a high retention rate (94%, [16/17]). On average, posts were viewed by 91% of participants, liked by 35%, and commented on by 26%. The average comment rate was highest (65%) for posts that elicited comments by directly posing questions or discussion topics to the group. Average intervention acceptability ratings were highly positive and participants reported feeling connected to the group and its topic. Average rates of past 1-month indoor tanning reported following the intervention were lower than the baseline rate (P=.08, Cohen d=0.47). Conclusions This study is novel in demonstrating participant engagement with and acceptability of using Facebook secret groups to deliver a dissonance-inducing intervention approach that utilizes group-based discussions related to body image. The study is also unique within the field of skin cancer prevention by demonstrating the potential value of delivering an indoor tanning intervention within

  5. Effective Fingerprint Quality Estimation for Diverse Capture Sensors

    PubMed Central

    Xie, Shan Juan; Yoon, Sook; Shin, Jinwook; Park, Dong Sun

    2010-01-01

    Recognizing the quality of fingerprints in advance can be beneficial for improving the performance of fingerprint recognition systems. The representative features to assess the quality of fingerprint images from different types of capture sensors are known to vary. In this paper, an effective quality estimation system that can be adapted for different types of capture sensors is designed by modifying and combining a set of features including orientation certainty, local orientation quality and consistency. The proposed system extracts basic features, and generates next level features which are applicable for various types of capture sensors. The system then uses the Support Vector Machine (SVM) classifier to determine whether or not an image should be accepted as input to the recognition system. The experimental results show that the proposed method can perform better than previous methods in terms of accuracy. In the meanwhile, the proposed method has an ability to eliminate residue images from the optical and capacitive sensors, and the coarse images from thermal sensors. PMID:22163632

  6. Quality assessment of color images based on the measure of just noticeable color difference

    NASA Astrophysics Data System (ADS)

    Chou, Chun-Hsien; Hsu, Yun-Hsiang

    2014-01-01

    Accurate assessment on the quality of color images is an important step to many image processing systems that convey visual information of the reproduced images. An accurate objective image quality assessment (IQA) method is expected to give the assessment result highly agreeing with the subjective assessment. To assess the quality of color images, many approaches simply apply the metric for assessing the quality of gray scale images to each of three color channels of the color image, neglecting the correlation among three color channels. In this paper, a metric for assessing color images' quality is proposed, in which the model of variable just-noticeable color difference (VJNCD) is employed to estimate the visibility thresholds of distortion inherent in each color pixel. With the estimated visibility thresholds of distortion, the proposed metric measures the average perceptible distortion in terms of the quantized distortion according to the perceptual error map similar to that defined by National Bureau of Standards (NBS) for converting the color difference enumerated by CIEDE2000 to the objective score of perceptual quality assessment. The perceptual error map in this case is designed for each pixel according to the visibility threshold estimated by the VJNCD model. The performance of the proposed metric is verified by assessing the test images in the LIVE database, and is compared with those of many well-know IQA metrics. Experimental results indicate that the proposed metric is an effective IQA method that can accurately predict the image quality of color images in terms of the correlation between objective scores and subjective evaluation.

  7. Recent Developments in Hyperspectral Imaging for Assessment of Food Quality and Safety

    PubMed Central

    Huang, Hui; Liu, Li; Ngadi, Michael O.

    2014-01-01

    Hyperspectral imaging which combines imaging and spectroscopic technology is rapidly gaining ground as a non-destructive, real-time detection tool for food quality and safety assessment. Hyperspectral imaging could be used to simultaneously obtain large amounts of spatial and spectral information on the objects being studied. This paper provides a comprehensive review on the recent development of hyperspectral imaging applications in food and food products. The potential and future work of hyperspectral imaging for food quality and safety control is also discussed. PMID:24759119

  8. A cross-platform survey of CT image quality and dose from routine abdomen protocols and a method to systematically standardize image quality.

    PubMed

    Favazza, Christopher P; Duan, Xinhui; Zhang, Yi; Yu, Lifeng; Leng, Shuai; Kofler, James M; Bruesewitz, Michael R; McCollough, Cynthia H

    2015-11-07

    Through this investigation we developed a methodology to evaluate and standardize CT image quality from routine abdomen protocols across different manufacturers and models. The influence of manufacturer-specific automated exposure control systems on image quality was directly assessed to standardize performance across a range of patient sizes. We evaluated 16 CT scanners across our health system, including Siemens, GE, and Toshiba models. Using each practice's routine abdomen protocol, we measured spatial resolution, image noise, and scanner radiation output (CTDIvol). Axial and in-plane spatial resolutions were assessed through slice sensitivity profile (SSP) and modulation transfer function (MTF) measurements, respectively. Image noise and CTDIvol values were obtained for three different phantom sizes. SSP measurements demonstrated a bimodal distribution in slice widths: an average of 6.2  ±  0.2 mm using GE's 'Plus' mode reconstruction setting and 5.0  ±  0.1 mm for all other scanners. MTF curves were similar for all scanners. Average spatial frequencies at 50%, 10%, and 2% MTF values were 3.24  ±  0.37, 6.20  ±  0.34, and 7.84  ±  0.70 lp cm(-1), respectively. For all phantom sizes, image noise and CTDIvol varied considerably: 6.5-13.3 HU (noise) and 4.8-13.3 mGy (CTDIvol) for the smallest phantom; 9.1-18.4 HU and 9.3-28.8 mGy for the medium phantom; and 7.8-23.4 HU and 16.0-48.1 mGy for the largest phantom. Using these measurements and benchmark SSP, MTF, and image noise targets, CT image quality can be standardized across a range of patient sizes.

  9. Investigation of cone-beam CT image quality trade-off for image-guided radiation therapy

    NASA Astrophysics Data System (ADS)

    Bian, Junguo; Sharp, Gregory C.; Park, Yang-Kyun; Ouyang, Jinsong; Bortfeld, Thomas; El Fakhri, Georges

    2016-05-01

    It is well-known that projections acquired over an angular range slightly over 180° (so-called short scan) are sufficient for fan-beam reconstruction. However, due to practical imaging conditions (projection data and reconstruction image discretization, physical factors, and data noise), the short-scan reconstructions may have different appearances and properties from the full-scan (scans over 360°) reconstructions. Nevertheless, short-scan configurations have been used in applications such as cone-beam CT (CBCT) for head-neck-cancer image-guided radiation therapy (IGRT) that only requires a small field of view due to the potential reduced imaging time and dose. In this work, we studied the image quality trade-off for full, short, and full/short scan configurations with both conventional filtered-backprojection (FBP) reconstruction and iterative reconstruction algorithms based on total-variation (TV) minimization for head-neck-cancer IGRT. Anthropomorphic and Catphan phantoms were scanned at different exposure levels with a clinical scanner used in IGRT. Both visualization- and numerical-metric-based evaluation studies were performed. The results indicate that the optimal exposure level and number of views are in the middle range for both FBP and TV-based iterative algorithms and the optimization is object-dependent and task-dependent. The optimal view numbers decrease with the total exposure levels for both FBP and TV-based algorithms. The results also indicate there are slight differences between FBP and TV-based iterative algorithms for the image quality trade-off: FBP seems to be more in favor of larger number of views while the TV-based algorithm is more robust to different data conditions (number of views and exposure levels) than the FBP algorithm. The studies can provide a general guideline for image-quality optimization for CBCT used in IGRT and other applications.

  10. Investigation of cone-beam CT image quality trade-off for image-guided radiation therapy.

    PubMed

    Bian, Junguo; Sharp, Gregory C; Park, Yang-Kyun; Ouyang, Jinsong; Bortfeld, Thomas; El Fakhri, Georges

    2016-05-07

    It is well-known that projections acquired over an angular range slightly over 180° (so-called short scan) are sufficient for fan-beam reconstruction. However, due to practical imaging conditions (projection data and reconstruction image discretization, physical factors, and data noise), the short-scan reconstructions may have different appearances and properties from the full-scan (scans over 360°) reconstructions. Nevertheless, short-scan configurations have been used in applications such as cone-beam CT (CBCT) for head-neck-cancer image-guided radiation therapy (IGRT) that only requires a small field of view due to the potential reduced imaging time and dose. In this work, we studied the image quality trade-off for full, short, and full/short scan configurations with both conventional filtered-backprojection (FBP) reconstruction and iterative reconstruction algorithms based on total-variation (TV) minimization for head-neck-cancer IGRT. Anthropomorphic and Catphan phantoms were scanned at different exposure levels with a clinical scanner used in IGRT. Both visualization- and numerical-metric-based evaluation studies were performed. The results indicate that the optimal exposure level and number of views are in the middle range for both FBP and TV-based iterative algorithms and the optimization is object-dependent and task-dependent. The optimal view numbers decrease with the total exposure levels for both FBP and TV-based algorithms. The results also indicate there are slight differences between FBP and TV-based iterative algorithms for the image quality trade-off: FBP seems to be more in favor of larger number of views while the TV-based algorithm is more robust to different data conditions (number of views and exposure levels) than the FBP algorithm. The studies can provide a general guideline for image-quality optimization for CBCT used in IGRT and other applications.

  11. A new method to evaluate image quality of CBCT images quantitatively without observers

    PubMed Central

    Shimizu, Mayumi; Okamura, Kazutoshi; Yoshida, Shoko; Weerawanich, Warangkana; Tokumori, Kenji; Jasa, Gainer R; Yoshiura, Kazunori

    2017-01-01

    Objectives: To develop an observer-free method for quantitatively evaluating the image quality of CBCT images by applying just-noticeable difference (JND). Methods: We used two test objects: (1) a Teflon (polytetrafluoroethylene) plate phantom attached to a dry human mandible; and (2) a block phantom consisting of a Teflon step phantom and an aluminium step phantom. These phantoms had holes with different depths. They were immersed in water and scanned with a CB MercuRay (Hitachi Medical Corporation, Tokyo, Japan) at tube voltages of 120 kV, 100 kV, 80 kV and 60 kV. Superimposed images of the phantoms with holes were used for evaluation. The number of detectable holes was used as an index of image quality. In detecting holes quantitatively, the threshold grey value (ΔG), which differentiated holes from the background, was calculated using a specific threshold (the JND), and we extracted the holes with grey values above ΔG. The indices obtained by this quantitative method (the extracted hole values) were compared with the observer evaluations (the observed hole values). In addition, the contrast-to-noise ratio (CNR) of the shallowest detectable holes and the deepest undetectable holes were measured to evaluate the contribution of CNR to detectability. Results: The results of this evaluation method corresponded almost exactly with the evaluations made by observers. The extracted hole values reflected the influence of different tube voltages. All extracted holes had an area with a CNR of ≥1.5. Conclusions: This quantitative method of evaluating CBCT image quality may be more useful and less time-consuming than evaluation by observation. PMID:28045343

  12. Full-reference quality assessment of stereoscopic images by learning binocular receptive field properties.

    PubMed

    Shao, Feng; Li, Kemeng; Lin, Weisi; Jiang, Gangyi; Yu, Mei; Dai, Qionghai

    2015-10-01

    Quality assessment of 3D images encounters more challenges than its 2D counterparts. Directly applying 2D image quality metrics is not the solution. In this paper, we propose a new full-reference quality assessment for stereoscopic images by learning binocular receptive field properties to be more in line with human visual perception. To be more specific, in the training phase, we learn a multiscale dictionary from the training database, so that the latent structure of images can be represented as a set of basis vectors. In the quality estimation phase, we compute sparse feature similarity index based on the estimated sparse coefficient vectors by considering their phase difference and amplitude difference, and compute global luminance similarity index by considering luminance changes. The final quality score is obtained by incorporating binocular combination based on sparse energy and sparse complexity. Experimental results on five public 3D image quality assessment databases demonstrate that in comparison with the most related existing methods, the devised algorithm achieves high consistency with subjective assessment.

  13. Deep architecture neural network-based real-time image processing for image-guided radiotherapy.

    PubMed

    Mori, Shinichiro

    2017-08-01

    To develop real-time image processing for image-guided radiotherapy, we evaluated several neural network models for use with different imaging modalities, including X-ray fluoroscopic image denoising. Setup images of prostate cancer patients were acquired with two oblique X-ray fluoroscopic units. Two types of residual network were designed: a convolutional autoencoder (rCAE) and a convolutional neural network (rCNN). We changed the convolutional kernel size and number of convolutional layers for both networks, and the number of pooling and upsampling layers for rCAE. The ground-truth image was applied to the contrast-limited adaptive histogram equalization (CLAHE) method of image processing. Network models were trained to keep the quality of the output image close to that of the ground-truth image from the input image without image processing. For image denoising evaluation, noisy input images were used for the training. More than 6 convolutional layers with convolutional kernels >5×5 improved image quality. However, this did not allow real-time imaging. After applying a pair of pooling and upsampling layers to both networks, rCAEs with >3 convolutions each and rCNNs with >12 convolutions with a pair of pooling and upsampling layers achieved real-time processing at 30 frames per second (fps) with acceptable image quality. Use of our suggested network achieved real-time image processing for contrast enhancement and image denoising by the use of a conventional modern personal computer. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  14. Study of quality perception in medical images based on comparison of contrast enhancement techniques in mammographic images

    NASA Astrophysics Data System (ADS)

    Matheus, B.; Verçosa, L. B.; Barufaldi, B.; Schiabel, H.

    2014-03-01

    With the absolute prevalence of digital images in mammography several new tools became available for radiologist; such as CAD schemes, digital zoom and contrast alteration. This work focuses in contrast variation and how the radiologist reacts to these changes when asked to evaluated image quality. Three contrast enhancing techniques were used in this study: conventional equalization, CCB Correction [1] - a digitization correction - and value subtraction. A set of 100 images was used in tests from some available online mammographic databases. The tests consisted of the presentation of all four versions of an image (original plus the three contrast enhanced images) to the specialist, requested to rank each one from the best up to worst quality for diagnosis. Analysis of results has demonstrated that CCB Correction [1] produced better images in almost all cases. Equalization, which mathematically produces a better contrast, was considered the worst for mammography image quality enhancement in the majority of cases (69.7%). The value subtraction procedure produced images considered better than the original in 84% of cases. Tests indicate that, for the radiologist's perception, it seems more important to guaranty full visualization of nuances than a high contrast image. Another result observed is that the "ideal" scanner curve does not yield the best result for a mammographic image. The important contrast range is the middle of the histogram, where nodules and masses need to be seen and clearly distinguished.

  15. Acceptance Probability (P a) Analysis for Process Validation Lifecycle Stages.

    PubMed

    Alsmeyer, Daniel; Pazhayattil, Ajay; Chen, Shu; Munaretto, Francesco; Hye, Maksuda; Sanghvi, Pradeep

    2016-04-01

    This paper introduces an innovative statistical approach towards understanding how variation impacts the acceptance criteria of quality attributes. Because of more complex stage-wise acceptance criteria, traditional process capability measures are inadequate for general application in the pharmaceutical industry. The probability of acceptance concept provides a clear measure, derived from specific acceptance criteria for each quality attribute. In line with the 2011 FDA Guidance, this approach systematically evaluates data and scientifically establishes evidence that a process is capable of consistently delivering quality product. The probability of acceptance provides a direct and readily understandable indication of product risk. As with traditional capability indices, the acceptance probability approach assumes that underlying data distributions are normal. The computational solutions for dosage uniformity and dissolution acceptance criteria are readily applicable. For dosage uniformity, the expected AV range may be determined using the s lo and s hi values along with the worst case estimates of the mean. This approach permits a risk-based assessment of future batch performance of the critical quality attributes. The concept is also readily applicable to sterile/non sterile liquid dose products. Quality attributes such as deliverable volume and assay per spray have stage-wise acceptance that can be converted into an acceptance probability. Accepted statistical guidelines indicate processes with C pk > 1.33 as performing well within statistical control and those with C pk < 1.0 as "incapable" (1). A C pk > 1.33 is associated with a centered process that will statistically produce less than 63 defective units per million. This is equivalent to an acceptance probability of >99.99%.

  16. Study of Image Qualities From 6D Robot-Based CBCT Imaging System of Small Animal Irradiator.

    PubMed

    Sharma, Sunil; Narayanasamy, Ganesh; Clarkson, Richard; Chao, Ming; Moros, Eduardo G; Zhang, Xin; Yan, Yulong; Boerma, Marjan; Paudel, Nava; Morrill, Steven; Corry, Peter; Griffin, Robert J

    2017-01-01

    To assess the quality of cone beam computed tomography images obtained by a robotic arm-based and image-guided small animal conformal radiation therapy device. The small animal conformal radiation therapy device is equipped with a 40 to 225 kV X-ray tube mounted on a custom made gantry, a 1024 × 1024 pixels flat panel detector (200 μm resolution), a programmable 6 degrees of freedom robot for cone beam computed tomography imaging and conformal delivery of radiation doses. A series of 2-dimensional radiographic projection images were recorded in cone beam mode by placing and rotating microcomputed tomography phantoms on the "palm' of the robotic arm. Reconstructed images were studied for image quality (spatial resolution, image uniformity, computed tomography number linearity, voxel noise, and artifacts). Geometric accuracy was measured to be 2% corresponding to 0.7 mm accuracy on a Shelley microcomputed tomography QA phantom. Qualitative resolution of reconstructed axial computed tomography slices using the resolution coils was within 200 μm. Quantitative spatial resolution was found to be 3.16 lp/mm. Uniformity of the system was measured within 34 Hounsfield unit on a QRM microcomputed tomography water phantom. Computed tomography numbers measured using the linearity plate were linear with material density ( R 2 > 0.995). Cone beam computed tomography images of the QRM multidisk phantom had minimal artifacts. Results showed that the small animal conformal radiation therapy device is capable of producing high-quality cone beam computed tomography images for precise and conformal small animal dose delivery. With its high-caliber imaging capabilities, the small animal conformal radiation therapy device is a powerful tool for small animal research.

  17. Evaluation of the sparse coding super-resolution method for improving image quality of up-sampled images in computed tomography

    NASA Astrophysics Data System (ADS)

    Ota, Junko; Umehara, Kensuke; Ishimaru, Naoki; Ohno, Shunsuke; Okamoto, Kentaro; Suzuki, Takanori; Shirai, Naoki; Ishida, Takayuki

    2017-02-01

    As the capability of high-resolution displays grows, high-resolution images are often required in Computed Tomography (CT). However, acquiring high-resolution images takes a higher radiation dose and a longer scanning time. In this study, we applied the Sparse-coding-based Super-Resolution (ScSR) method to generate high-resolution images without increasing the radiation dose. We prepared the over-complete dictionary learned the mapping between low- and highresolution patches and seek a sparse representation of each patch of the low-resolution input. These coefficients were used to generate the high-resolution output. For evaluation, 44 CT cases were used as the test dataset. We up-sampled images up to 2 or 4 times and compared the image quality of the ScSR scheme and bilinear and bicubic interpolations, which are the traditional interpolation schemes. We also compared the image quality of three learning datasets. A total of 45 CT images, 91 non-medical images, and 93 chest radiographs were used for dictionary preparation respectively. The image quality was evaluated by measuring peak signal-to-noise ratio (PSNR) and structure similarity (SSIM). The differences of PSNRs and SSIMs between the ScSR method and interpolation methods were statistically significant. Visual assessment confirmed that the ScSR method generated a high-resolution image with sharpness, whereas conventional interpolation methods generated over-smoothed images. To compare three different training datasets, there were no significance between the CT, the CXR and non-medical datasets. These results suggest that the ScSR provides a robust approach for application of up-sampling CT images and yields substantial high image quality of extended images in CT.

  18. Dried fruits quality assessment by hyperspectral imaging

    NASA Astrophysics Data System (ADS)

    Serranti, Silvia; Gargiulo, Aldo; Bonifazi, Giuseppe

    2012-05-01

    Dried fruits products present different market values according to their quality. Such a quality is usually quantified in terms of freshness of the products, as well as presence of contaminants (pieces of shell, husk, and small stones), defects, mould and decays. The combination of these parameters, in terms of relative presence, represent a fundamental set of attributes conditioning dried fruits humans-senses-detectable-attributes (visual appearance, organolectic properties, etc.) and their overall quality in terms of marketable products. Sorting-selection strategies exist but sometimes they fail when a higher degree of detection is required especially if addressed to discriminate between dried fruits of relatively small dimensions and when aiming to perform an "early detection" of pathogen agents responsible of future moulds and decays development. Surface characteristics of dried fruits can be investigated by hyperspectral imaging (HSI). In this paper, specific and "ad hoc" applications addressed to propose quality detection logics, adopting a hyperspectral imaging (HSI) based approach, are described, compared and critically evaluated. Reflectance spectra of selected dried fruits (hazelnuts) of different quality and characterized by the presence of different contaminants and defects have been acquired by a laboratory device equipped with two HSI systems working in two different spectral ranges: visible-near infrared field (400-1000 nm) and near infrared field (1000-1700 nm). The spectra have been processed and results evaluated adopting both a simple and fast wavelength band ratio approach and a more sophisticated classification logic based on principal component (PCA) analysis.

  19. Standardisation of DNA quantitation by image analysis: quality control of instrumentation.

    PubMed

    Puech, M; Giroud, F

    1999-05-01

    DNA image analysis is frequently performed in clinical practice as a prognostic tool and to improve diagnosis. The precision of prognosis and diagnosis depends on the accuracy of analysis and particularly on the quality of image analysis systems. It has been reported that image analysis systems used for DNA quantification differ widely in their characteristics (Thunissen et al.: Cytometry 27: 21-25, 1997). This induces inter-laboratory variations when the same sample is analysed in different laboratories. In microscopic image analysis, the principal instrumentation errors arise from the optical and electronic parts of systems. They bring about problems of instability, non-linearity, and shading and glare phenomena. The aim of this study is to establish tools and standardised quality control procedures for microscopic image analysis systems. Specific reference standard slides have been developed to control instability, non-linearity, shading and glare phenomena and segmentation efficiency. Some systems have been controlled with these tools and these quality control procedures. Interpretation criteria and accuracy limits of these quality control procedures are proposed according to the conclusions of a European project called PRESS project (Prototype Reference Standard Slide). Beyond these limits, tested image analysis systems are not qualified to realise precise DNA analysis. The different procedures presented in this work determine if an image analysis system is qualified to deliver sufficiently precise DNA measurements for cancer case analysis. If the controlled systems are beyond the defined limits, some recommendations are given to find a solution to the problem.

  20. Content-based quality evaluation of color images: overview and proposals

    NASA Astrophysics Data System (ADS)

    Tremeau, Alain; Richard, Noel; Colantoni, Philippe; Fernandez-Maloigne, Christine

    2003-12-01

    The automatic prediction of perceived quality from image data in general, and the assessment of particular image characteristics or attributes that may need improvement in particular, becomes an increasingly important part of intelligent imaging systems. The purpose of this paper is to propose to the color imaging community in general to develop a software package available on internet to help the user to select among all these approaches which is better appropriated to a given application. The ultimate goal of this project is to propose, next to implement, an open and unified color imaging system to set up a favourable context for the evaluation and analysis of color imaging processes. Many different methods for measuring the performance of a process have been proposed by different researchers. In this paper, we will discuss the advantages and shortcomings of most of main analysis criteria and performance measures currently used. The aim is not to establish a harsh competition between algorithms or processes, but rather to test and compare the efficiency of methodologies firstly to highlight strengths and weaknesses of a given algorithm or methodology on a given image type and secondly to have these results publicly available. This paper is focused on two important unsolved problems. Why it is so difficult to select a color space which gives better results than another one? Why it is so difficult to select an image quality metric which gives better results than another one, with respect to the judgment of the Human Visual System? Several methods used either in color imaging or in image quality will be thus discussed. Proposals for content-based image measures and means of developing a standard test suite for will be then presented. The above reference advocates for an evaluation protocol based on an automated procedure. This is the ultimate goal of our proposal.

  1. Spatially-controlled illumination with rescan confocal microscopy enhances image quality, resolution and reduces photodamage

    NASA Astrophysics Data System (ADS)

    Krishnaswami, Venkataraman; De Luca, Giulia M. R.; Breedijk, Ronald M. P.; Van Noorden, Cornelis J. F.; Manders, Erik M. M.; Hoebe, Ron A.

    2017-02-01

    Fluorescence microscopy is an important tool in biomedical imaging. An inherent trade-off lies between image quality and photodamage. Recently, we have introduced rescan confocal microscopy (RCM) that improves the lateral resolution of a confocal microscope down to 170 nm. Previously, we have demonstrated that with controlled-light exposure microscopy, spatial control of illumination reduces photodamage without compromising image quality. Here, we show that the combination of these two techniques leads to high resolution imaging with reduced photodamage without compromising image quality. Implementation of spatially-controlled illumination was carried out in RCM using a line scanning-based approach. Illumination is spatially-controlled for every line during imaging with the help of a prediction algorithm that estimates the spatial profile of the fluorescent specimen. The estimation is based on the information available from previously acquired line images. As a proof-of-principle, we show images of N1E-115 neuroblastoma cells, obtained by this new setup with reduced illumination dose, improved resolution and without compromising image quality.

  2. Restoration of color in a remote sensing image and its quality evaluation

    NASA Astrophysics Data System (ADS)

    Zhang, Zuxun; Li, Zhijiang; Zhang, Jianqing; Wang, Zhihe

    2003-09-01

    This paper is focused on the restoration of color remote sensing (including airborne photo). A complete approach is recommended. It propose that two main aspects should be concerned in restoring a remote sensing image, that are restoration of space information, restoration of photometric information. In this proposal, the restoration of space information can be performed by making the modulation transfer function (MTF) as degradation function, in which the MTF is obtained by measuring the edge curve of origin image. The restoration of photometric information can be performed by improved local maximum entropy algorithm. What's more, a valid approach in processing color remote sensing image is recommended. That is splits the color remote sensing image into three monochromatic images which corresponding three visible light bands and synthesizes the three images after being processed separately with psychological color vision restriction. Finally, three novel evaluation variables are obtained based on image restoration to evaluate the image restoration quality in space restoration quality and photometric restoration quality. An evaluation is provided at last.

  3. Simultaneous Analysis and Quality Assurance for Diffusion Tensor Imaging

    PubMed Central

    Lauzon, Carolyn B.; Asman, Andrew J.; Esparza, Michael L.; Burns, Scott S.; Fan, Qiuyun; Gao, Yurui; Anderson, Adam W.; Davis, Nicole; Cutting, Laurie E.; Landman, Bennett A.

    2013-01-01

    Diffusion tensor imaging (DTI) enables non-invasive, cyto-architectural mapping of in vivo tissue microarchitecture through voxel-wise mathematical modeling of multiple magnetic resonance imaging (MRI) acquisitions, each differently sensitized to water diffusion. DTI computations are fundamentally estimation processes and are sensitive to noise and artifacts. Despite widespread adoption in the neuroimaging community, maintaining consistent DTI data quality remains challenging given the propensity for patient motion, artifacts associated with fast imaging techniques, and the possibility of hardware changes/failures. Furthermore, the quantity of data acquired per voxel, the non-linear estimation process, and numerous potential use cases complicate traditional visual data inspection approaches. Currently, quality inspection of DTI data has relied on visual inspection and individual processing in DTI analysis software programs (e.g. DTIPrep, DTI-studio). However, recent advances in applied statistical methods have yielded several different metrics to assess noise level, artifact propensity, quality of tensor fit, variance of estimated measures, and bias in estimated measures. To date, these metrics have been largely studied in isolation. Herein, we select complementary metrics for integration into an automatic DTI analysis and quality assurance pipeline. The pipeline completes in 24 hours, stores statistical outputs, and produces a graphical summary quality analysis (QA) report. We assess the utility of this streamlined approach for empirical quality assessment on 608 DTI datasets from pediatric neuroimaging studies. The efficiency and accuracy of quality analysis using the proposed pipeline is compared with quality analysis based on visual inspection. The unified pipeline is found to save a statistically significant amount of time (over 70%) while improving the consistency of QA between a DTI expert and a pool of research associates. Projection of QA metrics to a low

  4. Simultaneous analysis and quality assurance for diffusion tensor imaging.

    PubMed

    Lauzon, Carolyn B; Asman, Andrew J; Esparza, Michael L; Burns, Scott S; Fan, Qiuyun; Gao, Yurui; Anderson, Adam W; Davis, Nicole; Cutting, Laurie E; Landman, Bennett A

    2013-01-01

    Diffusion tensor imaging (DTI) enables non-invasive, cyto-architectural mapping of in vivo tissue microarchitecture through voxel-wise mathematical modeling of multiple magnetic resonance imaging (MRI) acquisitions, each differently sensitized to water diffusion. DTI computations are fundamentally estimation processes and are sensitive to noise and artifacts. Despite widespread adoption in the neuroimaging community, maintaining consistent DTI data quality remains challenging given the propensity for patient motion, artifacts associated with fast imaging techniques, and the possibility of hardware changes/failures. Furthermore, the quantity of data acquired per voxel, the non-linear estimation process, and numerous potential use cases complicate traditional visual data inspection approaches. Currently, quality inspection of DTI data has relied on visual inspection and individual processing in DTI analysis software programs (e.g. DTIPrep, DTI-studio). However, recent advances in applied statistical methods have yielded several different metrics to assess noise level, artifact propensity, quality of tensor fit, variance of estimated measures, and bias in estimated measures. To date, these metrics have been largely studied in isolation. Herein, we select complementary metrics for integration into an automatic DTI analysis and quality assurance pipeline. The pipeline completes in 24 hours, stores statistical outputs, and produces a graphical summary quality analysis (QA) report. We assess the utility of this streamlined approach for empirical quality assessment on 608 DTI datasets from pediatric neuroimaging studies. The efficiency and accuracy of quality analysis using the proposed pipeline is compared with quality analysis based on visual inspection. The unified pipeline is found to save a statistically significant amount of time (over 70%) while improving the consistency of QA between a DTI expert and a pool of research associates. Projection of QA metrics to a low

  5. Image quality of multiplanar reconstruction of pulmonary CT scans using adaptive statistical iterative reconstruction.

    PubMed

    Honda, O; Yanagawa, M; Inoue, A; Kikuyama, A; Yoshida, S; Sumikawa, H; Tobino, K; Koyama, M; Tomiyama, N

    2011-04-01

    We investigated the image quality of multiplanar reconstruction (MPR) using adaptive statistical iterative reconstruction (ASIR). Inflated and fixed lungs were scanned with a garnet detector CT in high-resolution mode (HR mode) or non-high-resolution (HR) mode, and MPR images were then reconstructed. Observers compared 15 MPR images of ASIR (40%) and ASIR (80%) with those of ASIR (0%), and assessed image quality using a visual five-point scale (1, definitely inferior; 5, definitely superior), with particular emphasis on normal pulmonary structures, artefacts, noise and overall image quality. The mean overall image quality scores in HR mode were 3.67 with ASIR (40%) and 4.97 with ASIR (80%). Those in non-HR mode were 3.27 with ASIR (40%) and 3.90 with ASIR (80%). The mean artefact scores in HR mode were 3.13 with ASIR (40%) and 3.63 with ASIR (80%), but those in non-HR mode were 2.87 with ASIR (40%) and 2.53 with ASIR (80%). The mean scores of the other parameters were greater than 3, whereas those in HR mode were higher than those in non-HR mode. There were significant differences between ASIR (40%) and ASIR (80%) in overall image quality (p<0.01). Contrast medium in the injection syringe was scanned to analyse image quality; ASIR did not suppress the severe artefacts of contrast medium. In general, MPR image quality with ASIR (80%) was superior to that with ASIR (40%). However, there was an increased incidence of artefacts by ASIR when CT images were obtained in non-HR mode.

  6. Paediatric cerebrovascular CT angiography—towards better image quality

    PubMed Central

    Thust, Stefanie C.; Chong, Wui Khean Kling; Gunny, Roxana; Mazumder, Asif; Poitelea, Marius; Welsh, Anna; Ederies, Ash

    2014-01-01

    Background Paediatric cerebrovascular CT angiography (CTA) can be challenging to perform due to variable cardiovascular physiology between different age groups and the risk of movement artefact. This analysis aimed to determine what proportion of CTA at our institution was of diagnostic quality and identify technical factors which could be improved. Materials and methods a retrospective analysis of 20 cases was performed at a national paediatric neurovascular centre assessing image quality with a subjective scoring system and Hounsfield Unit (HU) measurements. Demographic data, contrast dose, flow rate and triggering times were recorded for each patient. Results Using a qualitative scoring system, 75% of studies were found to be of diagnostic quality (n=9 ‘good’, n=6 ‘satisfactory’) and 25% (n=5) were ‘poor’. Those judged subjectively to be poor had arterial contrast density measured at less than 250 HU. Increased arterial opacification was achieved for cases performed with an increased flow rate (2.5-4 mL/s) and higher intravenous contrast dose (2 mL/kg). Triggering was found to be well timed in nine cases, early in four cases and late in seven cases. Of the scans triggered early, 75% were poor. Of the scans triggered late, less (29%) were poor. Conclusions High flow rates (>2.5 mL/s) were a key factor for achieving high quality paediatric cerebrovascular CTA imaging. However, appropriate triggering by starting the scan immediately on contrast opacification of the monitoring vessel plays an important role and could maintain image quality when flow rates were lower. Early triggering appeared more detrimental than late. PMID:25525579

  7. Use of focus measure operators for characterization of flood illumination adaptive optics ophthalmoscopy image quality

    PubMed Central

    Alonso-Caneiro, David; Sampson, Danuta M.; Chew, Avenell L.; Collins, Michael J.; Chen, Fred K.

    2018-01-01

    Adaptive optics flood illumination ophthalmoscopy (AO-FIO) allows imaging of the cone photoreceptor in the living human retina. However, clinical interpretation of the AO-FIO image remains challenging due to suboptimal quality arising from residual uncorrected wavefront aberrations and rapid eye motion. An objective method of assessing image quality is necessary to determine whether an AO-FIO image is suitable for grading and diagnostic purpose. In this work, we explore the use of focus measure operators as a surrogate measure of AO-FIO image quality. A set of operators are tested on data sets acquired at different focal depths and different retinal locations from healthy volunteers. Our results demonstrate differences in focus measure operator performance in quantifying AO-FIO image quality. Further, we discuss the potential application of the selected focus operators in (i) selection of the best quality AO-FIO image from a series of images collected at the same retinal location and (ii) assessment of longitudinal changes in the diseased retina. Focus function could be incorporated into real-time AO-FIO image processing and provide an initial automated quality assessment during image acquisition or reading center grading. PMID:29552404

  8. Use of focus measure operators for characterization of flood illumination adaptive optics ophthalmoscopy image quality.

    PubMed

    Alonso-Caneiro, David; Sampson, Danuta M; Chew, Avenell L; Collins, Michael J; Chen, Fred K

    2018-02-01

    Adaptive optics flood illumination ophthalmoscopy (AO-FIO) allows imaging of the cone photoreceptor in the living human retina. However, clinical interpretation of the AO-FIO image remains challenging due to suboptimal quality arising from residual uncorrected wavefront aberrations and rapid eye motion. An objective method of assessing image quality is necessary to determine whether an AO-FIO image is suitable for grading and diagnostic purpose. In this work, we explore the use of focus measure operators as a surrogate measure of AO-FIO image quality. A set of operators are tested on data sets acquired at different focal depths and different retinal locations from healthy volunteers. Our results demonstrate differences in focus measure operator performance in quantifying AO-FIO image quality. Further, we discuss the potential application of the selected focus operators in (i) selection of the best quality AO-FIO image from a series of images collected at the same retinal location and (ii) assessment of longitudinal changes in the diseased retina. Focus function could be incorporated into real-time AO-FIO image processing and provide an initial automated quality assessment during image acquisition or reading center grading.

  9. TL dosimetry for quality control of CR mammography imaging systems

    NASA Astrophysics Data System (ADS)

    Gaona, E.; Nieto, J. A.; Góngora, J. A. I. D.; Arreola, M.; Enríquez, J. G. F.

    The aim of this work is to estimate the average glandular dose with thermoluminescent (TL) dosimetry and comparison with quality imaging in computed radiography (CR) mammography. For a measuring dose, the Food and Drug Administration (FDA) and the American College of Radiology (ACR) use a phantom, so that dose and image quality are assessed with the same test object. The mammography is a radiological image to visualize early biological manifestations of breast cancer. Digital systems have two types of image-capturing devices, full field digital mammography (FFDM) and CR mammography. In Mexico, there are several CR mammography systems in clinical use, but only one system has been approved for use by the FDA. Mammography CR uses a photostimulable phosphor detector (PSP) system. Most CR plates are made of 85% BaFBr and 15% BaFI doped with europium (Eu) commonly called barium flourohalideE We carry out an exploratory survey of six CR mammography units from three different manufacturers and six dedicated X-ray mammography units with fully automatic exposure. The results show three CR mammography units (50%) have a dose greater than 3.0 mGy without demonstrating improved image quality. The differences between doses averages from TLD system and dosimeter with ionization chamber are less than 10%. TLD system is a good option for average glandular dose measurement for X-rays with a HVL (0.35-0.38 mmAl) and kVp (24-26) used in quality control procedures with ACR Mammography Accreditation Phantom.

  10. Digitized hand-wrist radiographs: comparison of subjective and software-derived image quality at various compression ratios.

    PubMed

    McCord, Layne K; Scarfe, William C; Naylor, Rachel H; Scheetz, James P; Silveira, Anibal; Gillespie, Kevin R

    2007-05-01

    The objectives of this study were to compare the effect of JPEG 2000 compression of hand-wrist radiographs on observer image quality qualitative assessment and to compare with a software-derived quantitative image quality index. Fifteen hand-wrist radiographs were digitized and saved as TIFF and JPEG 2000 images at 4 levels of compression (20:1, 40:1, 60:1, and 80:1). The images, including rereads, were viewed by 13 orthodontic residents who determined the image quality rating on a scale of 1 to 5. A quantitative analysis was also performed by using a readily available software based on the human visual system (Image Quality Measure Computer Program, version 6.2, Mitre, Bedford, Mass). ANOVA was used to determine the optimal compression level (P < or =.05). When we compared subjective indexes, JPEG compression greater than 60:1 significantly reduced image quality. When we used quantitative indexes, the JPEG 2000 images had lower quality at all compression ratios compared with the original TIFF images. There was excellent correlation (R2 >0.92) between qualitative and quantitative indexes. Image Quality Measure indexes are more sensitive than subjective image quality assessments in quantifying image degradation with compression. There is potential for this software-based quantitative method in determining the optimal compression ratio for any image without the use of subjective raters.

  11. Image processing system performance prediction and product quality evaluation

    NASA Technical Reports Server (NTRS)

    Stein, E. K.; Hammill, H. B. (Principal Investigator)

    1976-01-01

    The author has identified the following significant results. A new technique for image processing system performance prediction and product quality evaluation was developed. It was entirely objective, quantitative, and general, and should prove useful in system design and quality control. The technique and its application to determination of quality control procedures for the Earth Resources Technology Satellite NASA Data Processing Facility are described.

  12. Effect of masking phase-only holograms on the quality of reconstructed images.

    PubMed

    Deng, Yuanbo; Chu, Daping

    2016-04-20

    A phase-only hologram modulates the phase of the incident light and diffracts it efficiently with low energy loss because of the minimum absorption. Much research attention has been focused on how to generate phase-only holograms, and little work has been done to understand the effect and limitation of their partial implementation, possibly due to physical defects and constraints, in particular as in the practical situations where a phase-only hologram is confined or needs to be sliced or tiled. The present study simulates the effect of masking phase-only holograms on the quality of reconstructed images in three different scenarios with different filling factors, filling positions, and illumination intensity profiles. Quantitative analysis confirms that the width of the image point spread function becomes wider and the image quality decreases, as expected, when the filling factor decreases, and the image quality remains the same for different filling positions as well. The width of the image point spread function as derived from different filling factors shows a consistent behavior to that as measured directly from the reconstructed image, especially as the filling factor becomes small. Finally, mask profiles of different shapes and intensity distributions are shown to have more complicated effects on the image point spread function, which in turn affects the quality and textures of the reconstructed image.

  13. Designing image segmentation studies: Statistical power, sample size and reference standard quality.

    PubMed

    Gibson, Eli; Hu, Yipeng; Huisman, Henkjan J; Barratt, Dean C

    2017-12-01

    Segmentation algorithms are typically evaluated by comparison to an accepted reference standard. The cost of generating accurate reference standards for medical image segmentation can be substantial. Since the study cost and the likelihood of detecting a clinically meaningful difference in accuracy both depend on the size and on the quality of the study reference standard, balancing these trade-offs supports the efficient use of research resources. In this work, we derive a statistical power calculation that enables researchers to estimate the appropriate sample size to detect clinically meaningful differences in segmentation accuracy (i.e. the proportion of voxels matching the reference standard) between two algorithms. Furthermore, we derive a formula to relate reference standard errors to their effect on the sample sizes of studies using lower-quality (but potentially more affordable and practically available) reference standards. The accuracy of the derived sample size formula was estimated through Monte Carlo simulation, demonstrating, with 95% confidence, a predicted statistical power within 4% of simulated values across a range of model parameters. This corresponds to sample size errors of less than 4 subjects and errors in the detectable accuracy difference less than 0.6%. The applicability of the formula to real-world data was assessed using bootstrap resampling simulations for pairs of algorithms from the PROMISE12 prostate MR segmentation challenge data set. The model predicted the simulated power for the majority of algorithm pairs within 4% for simulated experiments using a high-quality reference standard and within 6% for simulated experiments using a low-quality reference standard. A case study, also based on the PROMISE12 data, illustrates using the formulae to evaluate whether to use a lower-quality reference standard in a prostate segmentation study. Copyright © 2017 The Authors. Published by Elsevier B.V. All rights reserved.

  14. Online hyperspectral imaging system for evaluating quality of agricultural products

    NASA Astrophysics Data System (ADS)

    Mo, Changyeun; Kim, Giyoung; Lim, Jongguk

    2017-06-01

    The consumption of fresh-cut agricultural produce in Korea has been growing. The browning of fresh-cut vegetables that occurs during storage and foreign substances such as worms and slugs are some of the main causes of consumers' concerns with respect to safety and hygiene. The purpose of this study is to develop an on-line system for evaluating quality of agricultural products using hyperspectral imaging technology. The online evaluation system with single visible-near infrared hyperspectral camera in the range of 400 nm to 1000 nm that can assess quality of both surfaces of agricultural products such as fresh-cut lettuce was designed. Algorithms to detect browning surface were developed for this system. The optimal wavebands for discriminating between browning and sound lettuce as well as between browning lettuce and the conveyor belt were investigated using the correlation analysis and the one-way analysis of variance method. The imaging algorithms to discriminate the browning lettuces were developed using the optimal wavebands. The ratio image (RI) algorithm of the 533 nm and 697 nm images (RI533/697) for abaxial surface lettuce and the ratio image algorithm (RI533/697) and subtraction image (SI) algorithm (SI538-697) for adaxial surface lettuce had the highest classification accuracies. The classification accuracy of browning and sound lettuce was 100.0% and above 96.0%, respectively, for the both surfaces. The overall results show that the online hyperspectral imaging system could potentially be used to assess quality of agricultural products.

  15. Evaluation of two minimal-preparation regimes for CT colonography: optimising image quality and patient acceptability

    PubMed Central

    Pollentine, A; Mortimer, A; Mccoubrie, P; Archer, L

    2012-01-01

    Objective To compare a 2 day bowel preparation regime of barium, iodine and a mild stimulant laxative with a 1 day iodine-only regime for CT colonography (CTC). Methods 100 consecutive patients underwent CTC. The first 50 patients (Regime 1) ingested 1 bisacodyl tablet twice a day 3 days before CTC and 1 dose each of 50 ml of barium and 20 ml of iodinated contrast per day starting 2 days before CTC. The second 50 patients (Regime 2) ingested 3 doses of iodinated contrast over 24 h prior to CTC. Volumes of residual stool and fluid, and the effectiveness of stool and fluid tagging, were graded according to methods established by Taylor et al (Taylor S, Slaker A, Burling D, Tam E, Greenhalgh R, Gartner L, et al. CT colonography: optimisation, diagnostic performance and patient acceptability of reduced-laxative regimens using barium-based faecal tagging. Eur Radiol 2008; 18: 32–42). A 3 day low-residue diet was taken by both cohorts. Questionnaires rating the side-effects and burden of the bowel preparation were compared to a control cohort of patients undergoing barium enema. Results The proportion of colons producing none/scattered stool (score 1) was 90.3% with Regime 1 and 65.0% with Regime 2 (p<0.005). Any residual stool was significantly better tagged with Regime 1 (score 5), with 91.7% of Regime 1 exhibiting optimum tagging vs 71.3% of Regime 2 (p<0.05). No significant differences in side-effects between the bowel preparation regimes for CTC were elicited. Bowel preparation for barium enema was tolerated significantly worse than both of the CTC bowel preparation regimes. Conclusion Regime 1, containing a 3 day preparation of a mild laxative, barium and iodine, produced a significantly better prepared colon, with no difference in patient acceptability. PMID:22422379

  16. Evaluation of two minimal-preparation regimes for CT colonography: optimising image quality and patient acceptability.

    PubMed

    Pollentine, A; Mortimer, A; McCoubrie, P; Archer, L

    2012-08-01

    To compare a 2 day bowel preparation regime of barium, iodine and a mild stimulant laxative with a 1 day iodine-only regime for CT colonography (CTC). 100 consecutive patients underwent CTC. The first 50 patients (Regime 1) ingested 1 bisacodyl tablet twice a day 3 days before CTC and 1 dose each of 50 ml of barium and 20 ml of iodinated contrast per day starting 2 days before CTC. The second 50 patients (Regime 2) ingested 3 doses of iodinated contrast over 24 h prior to CTC. Volumes of residual stool and fluid, and the effectiveness of stool and fluid tagging, were graded according to methods established by Taylor et al (Taylor S, Slaker A, Burling D, Tam E, Greenhalgh R, Gartner L, et al. CT colonography: optimisation, diagnostic performance and patient acceptability of reduced-laxative regimens using barium-based faecal tagging. Eur Radiol 2008; 18: 32-42). A 3 day low-residue diet was taken by both cohorts. Questionnaires rating the side-effects and burden of the bowel preparation were compared to a control cohort of patients undergoing barium enema. The proportion of colons producing none/scattered stool (score 1) was 90.3% with Regime 1 and 65.0% with Regime 2 (p<0.005). Any residual stool was significantly better tagged with Regime 1 (score 5), with 91.7% of Regime 1 exhibiting optimum tagging vs 71.3% of Regime 2 (p<0.05). No significant differences in side-effects between the bowel preparation regimes for CTC were elicited. Bowel preparation for barium enema was tolerated significantly worse than both of the CTC bowel preparation regimes. Regime 1, containing a 3 day preparation of a mild laxative, barium and iodine, produced a significantly better prepared colon, with no difference in patient acceptability.

  17. A quality assurance framework for the fully automated and objective evaluation of image quality in cone-beam computed tomography

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Steiding, Christian; Kolditz, Daniel; Kalender, Willi A., E-mail: willi.kalender@imp.uni-erlangen.de

    Purpose: Thousands of cone-beam computed tomography (CBCT) scanners for vascular, maxillofacial, neurological, and body imaging are in clinical use today, but there is no consensus on uniform acceptance and constancy testing for image quality (IQ) and dose yet. The authors developed a quality assurance (QA) framework for fully automated and time-efficient performance evaluation of these systems. In addition, the dependence of objective Fourier-based IQ metrics on direction and position in 3D volumes was investigated for CBCT. Methods: The authors designed a dedicated QA phantom 10 cm in length consisting of five compartments, each with a diameter of 10 cm, andmore » an optional extension ring 16 cm in diameter. A homogeneous section of water-equivalent material allows measuring CT value accuracy, image noise and uniformity, and multidimensional global and local noise power spectra (NPS). For the quantitative determination of 3D high-contrast spatial resolution, the modulation transfer function (MTF) of centrally and peripherally positioned aluminum spheres was computed from edge profiles. Additional in-plane and axial resolution patterns were used to assess resolution qualitatively. The characterization of low-contrast detectability as well as CT value linearity and artifact behavior was tested by utilizing sections with soft-tissue-equivalent and metallic inserts. For an automated QA procedure, a phantom detection algorithm was implemented. All tests used in the dedicated QA program were initially verified in simulation studies and experimentally confirmed on a clinical dental CBCT system. Results: The automated IQ evaluation of volume data sets of the dental CBCT system was achieved with the proposed phantom requiring only one scan for the determination of all desired parameters. Typically, less than 5 min were needed for phantom set-up, scanning, and data analysis. Quantitative evaluation of system performance over time by comparison to previous examinations

  18. Does body image perception relate to quality of life in middle-aged women?

    PubMed Central

    Medeiros de Morais, Maria Socorro; Vieira, Mariana Carmem Apolinário; Moreira, Mayle Andrade; da Câmara, Saionara Maria Aires; Campos Cavalcanti Maciel, Álvaro; Almeida, Maria das Graças

    2017-01-01

    Objective In Brazil, information about the influence of body image on the various life domains of women in menopausal transition is scarce. Thus, the objective of the study was to analyze the relationship between body image and quality of life in middle-aged Brazilian women. Methods This was a cross-sectional study of 250 women between 40 and 65 years old, living in Parnamirim/RN, Brazil, who were evaluated in relation to body image and quality of life. For body image, women were classified as: dissatisfied due to low weight, satisfied (with their body weight) and dissatisfied due to being overweight. Quality of life was assessed through a questionnaire in which higher values indicate higher quality of life. Multiple linear regression was performed to analyze the relationship between body image and quality of life, adjusted for covariates that presented p<0.20 in the bivariate analysis. Results The average age was 52.1 (± 5.6) years, 82% of the women reported being dissatisfied due to being overweight, and 4.4% were dissatisfied due to having low weight. After multiple linear regression analyzes, body image remained associated with health (p<0.001), emotional (p = 0.016), and sexual (p = 0.048) domains of quality of life, as well as total score of the questionnaire (p<0.001). Conclusion Women who reported being dissatisfied with their body image due to having low weight or overweight had worse quality of life in comparison to those who were satisfied (with their body weight). PMID:28926575

  19. Improved quality of intrafraction kilovoltage images by triggered readout of unexposed frames

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Poulsen, Per Rugaard, E-mail: per.poulsen@rm.dk; Jonassen, Johnny; Jensen, Carsten

    2015-11-15

    Purpose: The gantry-mounted kilovoltage (kV) imager of modern linear accelerators can be used for real-time tumor localization during radiation treatment delivery. However, the kV image quality often suffers from cross-scatter from the megavoltage (MV) treatment beam. This study investigates readout of unexposed kV frames as a means to improve the kV image quality in a series of experiments and a theoretical model of the observed image quality improvements. Methods: A series of fluoroscopic images were acquired of a solid water phantom with an embedded gold marker and an air cavity with and without simultaneous radiation of the phantom with amore » 6 MV beam delivered perpendicular to the kV beam with 300 and 600 monitor units per minute (MU/min). An in-house built device triggered readout of zero, one, or multiple unexposed frames between the kV exposures. The unexposed frames contained part of the MV scatter, consequently reducing the amount of MV scatter accumulated in the exposed frames. The image quality with and without unexposed frame readout was quantified as the contrast-to-noise ratio (CNR) of the gold marker and air cavity for a range of imaging frequencies from 1 to 15 Hz. To gain more insight into the observed CNR changes, the image lag of the kV imager was measured and used as input in a simple model that describes the CNR with unexposed frame readout in terms of the contrast, kV noise, and MV noise measured without readout of unexposed frames. Results: Without readout of unexposed kV frames, the quality of intratreatment kV images decreased dramatically with reduced kV frequencies due to MV scatter. The gold marker was only visible for imaging frequencies ≥3 Hz at 300 MU/min and ≥5 Hz for 600 MU/min. Visibility of the air cavity required even higher imaging frequencies. Readout of multiple unexposed frames ensured visibility of both structures at all imaging frequencies and a CNR that was independent of the kV frame rate. The image lag was 12

  20. Optical aberrations, retinal image quality and eye growth: Experimentation and modeling

    NASA Astrophysics Data System (ADS)

    Tian, Yibin

    2007-12-01

    Retinal image quality is important for normal eye growth. Optical aberrations are of interest for two reasons: first, they degrade retinal images; second, they might provide some cues to defocus. Higher than normal ocular aberrations have been previously associated with human myopia. However, these studies were cross-sectional in design, and only reported aberrations in terms of root mean square (RMS) errors of Zernike coefficients, a poor metric of optical quality. This dissertation presents results from investigations of ocular optical aberrations, retinal image quality and eye growth in chicks and humans. A number of techniques were utilized, including Shack-Hartmann aberrometry, high-frequency A-scan ultrasonography, ciliary nerve section (CNX), photorefractive keratectomy (PRK) as well as computer simulations and modeling. A technique to extract light scatter information from Shack-Hartmann images was also developed. The main findings of the dissertation are summarized below. In young chicks, most ocular aberrations decreased with growth in both normal and CNX eyes, and there were diurnal fluctuations in some aberrations. Modeling suggested active reduction in higher order aberrations (HOAs) during early development. Although CNX eyes manifested greater than normal HOAs, they showed near normal growth. Retinal image degradation varied greatly among individual eyes post-PRK in young chicks. Including light scatter information into analyses of retinal image quality better estimated the latter. Albino eyes showed more severe retinal image degradation than normal eyes, due to increased optical aberrations and light scatter, but their growth was similar to those of normal eyes, implying that they are relatively insensitive to retina image quality. Although the above results questioned the influence of optical aberrations on early ocular growth, some optical quality metrics, derived from optical aberrations data, could predict how much the eyes of young chicks

  1. Near-infrared hyperspectral imaging for quality analysis of agricultural and food products

    NASA Astrophysics Data System (ADS)

    Singh, C. B.; Jayas, D. S.; Paliwal, J.; White, N. D. G.

    2010-04-01

    Agricultural and food processing industries are always looking to implement real-time quality monitoring techniques as a part of good manufacturing practices (GMPs) to ensure high-quality and safety of their products. Near-infrared (NIR) hyperspectral imaging is gaining popularity as a powerful non-destructive tool for quality analysis of several agricultural and food products. This technique has the ability to analyse spectral data in a spatially resolved manner (i.e., each pixel in the image has its own spectrum) by applying both conventional image processing and chemometric tools used in spectral analyses. Hyperspectral imaging technique has demonstrated potential in detecting defects and contaminants in meats, fruits, cereals, and processed food products. This paper discusses the methodology of hyperspectral imaging in terms of hardware, software, calibration, data acquisition and compression, and development of prediction and classification algorithms and it presents a thorough review of the current applications of hyperspectral imaging in the analyses of agricultural and food products.

  2. Image quality improvement in MDCT cardiac imaging via SMART-RECON method

    NASA Astrophysics Data System (ADS)

    Li, Yinsheng; Cao, Ximiao; Xing, Zhanfeng; Sun, Xuguang; Hsieh, Jiang; Chen, Guang-Hong

    2017-03-01

    Coronary CT angiography (CCTA) is a challenging imaging task currently limited by the achievable temporal resolution of modern Multi-Detector CT (MDCT) scanners. In this paper, the recently proposed SMARTRECON method has been applied in MDCT-based CCTA imaging to improve the image quality without any prior knowledge of cardiac motion. After the prospective ECG-gated data acquisition from a short-scan angular span, the acquired data were sorted into several sub-sectors of view angles; each corresponds to a 1/4th of the short-scan angular range. Information of the cardiac motion was thus encoded into the data in each view angle sub-sector. The SMART-RECON algorithm was then applied to jointly reconstruct several image volumes, each of which is temporally consistent with the data acquired in the corresponding view angle sub-sector. Extensive numerical simulations were performed to validate the proposed technique and investigate the performance dependence.

  3. Simulation analysis of space remote sensing image quality degradation induced by satellite platform vibration

    NASA Astrophysics Data System (ADS)

    Yang, Feng; Zhang, Xiaofang; Huang, Yu; Hao, Weiwei; Guo, Baiwei

    2012-11-01

    Satellite platform vibration causes the image quality to be degraded, it is necessary to study its influence on image quality. The forms of Satellite platform vibration consist of linear vibration, sinusoidal vibration and random vibration. Based on Matlab & Zemax, the simulation system has been developed for simulating impact caused by satellite platform vibration on image quality. Dynamic Data Exchange is used for the communication between Matlab and Zemax. The data of sinusoidal vibration are produced by sinusoidal curve with specific amplitude and frequency. The data of random vibration are obtained by combining sinusoidal signals with 10Hz, 100Hz and 200Hz's frequency, 100, 12, 1.9's amplitude and white noise with zero mean value. Satellite platform vibration data which produced by Matlab are added to the optical system, and its point spread function can be obtained by Zemax. Blurred image can be gained by making the convolution of PSF and the original image. The definition of the original image and the blurred image are evaluated by using average gradient values of image gray. The impact caused by the sine and random vibration of six DOFs on the image quality are respectively simulated. The simulation result reveal that the decenter of X-, Y-, Z- direction and the tilt of Z-direction have a little effect on image quality, while the tilt of X-, Y- direction make image quality seriously degraded. Thus, it can be concluded that correcting the error of satellite platform vibration by FSM is a viable and effective way.

  4. Model-based quantification of image quality

    NASA Technical Reports Server (NTRS)

    Hazra, Rajeeb; Miller, Keith W.; Park, Stephen K.

    1989-01-01

    In 1982, Park and Schowengerdt published an end-to-end analysis of a digital imaging system quantifying three principal degradation components: (1) image blur - blurring caused by the acquisition system, (2) aliasing - caused by insufficient sampling, and (3) reconstruction blur - blurring caused by the imperfect interpolative reconstruction. This analysis, which measures degradation as the square of the radiometric error, includes the sample-scene phase as an explicit random parameter and characterizes the image degradation caused by imperfect acquisition and reconstruction together with the effects of undersampling and random sample-scene phases. In a recent paper Mitchell and Netravelli displayed the visual effects of the above mentioned degradations and presented subjective analysis about their relative importance in determining image quality. The primary aim of the research is to use the analysis of Park and Schowengerdt to correlate their mathematical criteria for measuring image degradations with subjective visual criteria. Insight gained from this research can be exploited in the end-to-end design of optical systems, so that system parameters (transfer functions of the acquisition and display systems) can be designed relative to each other, to obtain the best possible results using quantitative measurements.

  5. Effects of processing conditions on mammographic image quality.

    PubMed

    Braeuning, M P; Cooper, H W; O'Brien, S; Burns, C B; Washburn, D B; Schell, M J; Pisano, E D

    1999-08-01

    Any given mammographic film will exhibit changes in sensitometric response and image resolution as processing variables are altered. Developer type, immersion time, and temperature have been shown to affect the contrast of the mammographic image and thus lesion visibility. The authors evaluated the effect of altering processing variables, including film type, developer type, and immersion time, on the visibility of masses, fibrils, and speaks in a standard mammographic phantom. Images of a phantom obtained with two screen types (Kodak Min-R and Fuji) and five film types (Kodak Min-R M, Min-R E, Min-R H; Fuji UM-MA HC, and DuPont Microvision-C) were processed with five different developer chemicals (Autex SE, DuPont HSD, Kodak RP, Picker 3-7-90, and White Mountain) at four different immersion times (24, 30, 36, and 46 seconds). Processor chemical activity was monitored with sensitometric strips, and developer temperatures were continuously measured. The film images were reviewed by two board-certified radiologists and two physicists with expertise in mammography quality control and were scored based on the visibility of calcifications, masses, and fibrils. Although the differences in the absolute scores were not large, the Kodak Min-R M and Fuji films exhibited the highest scores, and images developed in White Mountain and Autex chemicals exhibited the highest scores. For any film, several processing chemicals may be used to produce images of similar quality. Extended processing may no longer be necessary.

  6. Image quality of multiplanar reconstruction of pulmonary CT scans using adaptive statistical iterative reconstruction

    PubMed Central

    Honda, O; Yanagawa, M; Inoue, A; Kikuyama, A; Yoshida, S; Sumikawa, H; Tobino, K; Koyama, M; Tomiyama, N

    2011-01-01

    Objective We investigated the image quality of multiplanar reconstruction (MPR) using adaptive statistical iterative reconstruction (ASIR). Methods Inflated and fixed lungs were scanned with a garnet detector CT in high-resolution mode (HR mode) or non-high-resolution (HR) mode, and MPR images were then reconstructed. Observers compared 15 MPR images of ASIR (40%) and ASIR (80%) with those of ASIR (0%), and assessed image quality using a visual five-point scale (1, definitely inferior; 5, definitely superior), with particular emphasis on normal pulmonary structures, artefacts, noise and overall image quality. Results The mean overall image quality scores in HR mode were 3.67 with ASIR (40%) and 4.97 with ASIR (80%). Those in non-HR mode were 3.27 with ASIR (40%) and 3.90 with ASIR (80%). The mean artefact scores in HR mode were 3.13 with ASIR (40%) and 3.63 with ASIR (80%), but those in non-HR mode were 2.87 with ASIR (40%) and 2.53 with ASIR (80%). The mean scores of the other parameters were greater than 3, whereas those in HR mode were higher than those in non-HR mode. There were significant differences between ASIR (40%) and ASIR (80%) in overall image quality (p<0.01). Contrast medium in the injection syringe was scanned to analyse image quality; ASIR did not suppress the severe artefacts of contrast medium. Conclusion In general, MPR image quality with ASIR (80%) was superior to that with ASIR (40%). However, there was an increased incidence of artefacts by ASIR when CT images were obtained in non-HR mode. PMID:21081572

  7. Visual quality analysis for images degraded by different types of noise

    NASA Astrophysics Data System (ADS)

    Ponomarenko, Nikolay N.; Lukin, Vladimir V.; Ieremeyev, Oleg I.; Egiazarian, Karen O.; Astola, Jaakko T.

    2013-02-01

    Modern visual quality metrics take into account different peculiarities of the Human Visual System (HVS). One of them is described by the Weber-Fechner law and deals with the different sensitivity to distortions in image fragments with different local mean values (intensity, brightness). We analyze how this property can be incorporated into a metric PSNRHVS- M. It is shown that some improvement of its performance can be provided. Then, visual quality of color images corrupted by three types of i.i.d. noise (pure additive, pure multiplicative, and signal dependent, Poisson) is analyzed. Experiments with a group of observers are carried out for distorted color images created on the basis of TID2008 database. Several modern HVS-metrics are considered. It is shown that even the best metrics are unable to assess visual quality of distorted images adequately enough. The reasons for this deal with the observer's attention to certain objects in the test images, i.e., with semantic aspects of vision, which are worth taking into account in design of HVS-metrics.

  8. Assessing image quality of low-cost laparoscopic box trainers: options for residents training at home.

    PubMed

    Kiely, Daniel J; Stephanson, Kirk; Ross, Sue

    2011-10-01

    Low-cost laparoscopic box trainers built using home computers and webcams may provide residents with a useful tool for practice at home. This study set out to evaluate the image quality of low-cost laparoscopic box trainers compared with a commercially available model. Five low-cost laparoscopic box trainers including the components listed were compared in random order to one commercially available box trainer: A (high-definition USB 2.0 webcam, PC laptop), B (Firewire webcam, Mac laptop), C (high-definition USB 2.0 webcam, Mac laptop), D (standard USB webcam, PC desktop), E (Firewire webcam, PC desktop), and F (the TRLCD03 3-DMEd Standard Minimally Invasive Training System). Participants observed still image quality and performed a peg transfer task using each box trainer. Participants rated still image quality, image quality with motion, and whether the box trainer had sufficient image quality to be useful for training. Sixteen residents in obstetrics and gynecology took part in the study. The box trainers showing no statistically significant difference from the commercially available model were A, B, C, D, and E for still image quality; A for image quality with motion; and A and B for usefulness of the simulator based on image quality. The cost of the box trainers A-E is approximately $100 to $160 each, not including a computer or laparoscopic instruments. Laparoscopic box trainers built from a high-definition USB 2.0 webcam with a PC (box trainer A) or from a Firewire webcam with a Mac (box trainer B) provide image quality comparable with a commercial standard.

  9. How do we watch images? A case of change detection and quality estimation

    NASA Astrophysics Data System (ADS)

    Radun, Jenni; Leisti, Tuomas; Virtanen, Toni; Nyman, Göte

    2012-01-01

    The most common tasks in subjective image estimation are change detection (a detection task) and image quality estimation (a preference task). We examined how the task influences the gaze behavior when comparing detection and preference tasks. The eye movements of 16 naïve observers were recorded with 8 observers in both tasks. The setting was a flicker paradigm, where the observers see a non-manipulated image, a manipulated version of the image and again the non-manipulated image and estimate the difference they perceived in them. The material was photographic material with different image distortions and contents. To examine the spatial distribution of fixations, we defined the regions of interest using a memory task and calculated information entropy to estimate how concentrated the fixations were on the image plane. The quality task was faster and needed fewer fixations and the first eight fixations were more concentrated on certain image areas than the change detection task. The bottom-up influences of the image also caused more variation to the gaze behavior in the quality estimation task than in the change detection task The results show that the quality estimation is faster and the regions of interest are emphasized more on certain images compared with the change detection task that is a scan task where the whole image is always thoroughly examined. In conclusion, in subjective image estimation studies it is important to think about the task.

  10. Quality assessment of butter cookies applying multispectral imaging

    PubMed Central

    Andresen, Mette S; Dissing, Bjørn S; Løje, Hanne

    2013-01-01

    A method for characterization of butter cookie quality by assessing the surface browning and water content using multispectral images is presented. Based on evaluations of the browning of butter cookies, cookies were manually divided into groups. From this categorization, reference values were calculated for a statistical prediction model correlating multispectral images with a browning score. The browning score is calculated as a function of oven temperature and baking time. It is presented as a quadratic response surface. The investigated process window was the intervals 4–16 min and 160–200°C in a forced convection electrically heated oven. In addition to the browning score, a model for predicting the average water content based on the same images is presented. This shows how multispectral images of butter cookies may be used for the assessment of different quality parameters. Statistical analysis showed that the most significant wavelengths for browning predictions were in the interval 400–700 nm and the wavelengths significant for water prediction were primarily located in the near-infrared spectrum. The water prediction model was found to correctly estimate the average water content with an absolute error of 0.22%. From the images it was also possible to follow the browning and drying propagation from the cookie edge toward the center. PMID:24804036

  11. Image quality and radiation dose on digital chest imaging: comparison of amorphous silicon and amorphous selenium flat-panel systems.

    PubMed

    Bacher, Klaus; Smeets, Peter; Vereecken, Ludo; De Hauwere, An; Duyck, Philippe; De Man, Robert; Verstraete, Koenraad; Thierens, Hubert

    2006-09-01

    The aim of this study was to compare the image quality and radiation dose in chest imaging using an amorphous silicon flat-panel detector system and an amorphous selenium flat-panel detector system. In addition, the low-contrast performance of both systems with standard and low radiation doses was compared. In two groups of 100 patients each, digital chest radiographs were acquired with either an amorphous silicon or an amorphous selenium flat-panel system. The effective dose of the examination was measured using thermoluminescent dosimeters placed in an anthropomorphic Rando phantom. The image quality of the digital chest radiographs was assessed by five experienced radiologists using the European Guidelines on Quality Criteria for Diagnostic Radiographic Images. In addition, a contrast-detail phantom study was set up to assess the low-contrast performance of both systems at different radiation dose levels. Differences between the two groups were tested for significance using the two-tailed Mann-Whitney test. The amorphous silicon flat-panel system allowed an important and significant reduction in effective dose in comparison with the amorphous selenium flat-panel system (p < 0.0001) for both the posteroanterior and lateral views. In addition, clinical image quality analysis showed that the dose reduction was not detrimental to image quality. Compared with the amorphous selenium flat-panel detector system, the amorphous silicon flat-panel detector system performed significantly better in the low-contrast phantom study, with phantom entrance dose values of up to 135 muGy. Chest radiographs can be acquired with a significantly lower patient radiation dose using an amorphous silicon flat-panel system than using an amorphous selenium flat-panel system, thereby producing images that are equal or even superior in quality to those of the amorphous selenium flat-panel detector system.

  12. Effect of exposure factors on image quality in screening mammography.

    PubMed

    Alkhalifah, K; Brindabhan, A; Alsaeed, R

    2017-11-01

    The aim of this research was to study the effect of exposure factors on image quality for digital screening mammography units in Kuwait which use Tungsten (W) targets with Rhodium (Rh) and Silver (Ag) as filters. Mammography Accreditation Phantom Model 015 was imaged using a Hologic Selenia Digital mammography unit with W targets and Rh and Ag filters. Four images, each at 26, 28, 30, and 32 kVp, were obtained using each target-filter combination (W/Rh and W/Ag). The images were evaluated by five senior technologists for the number of specks, fibers and masses visible on each image. Statistical analysis was carried out using non-parametric tests at p = 0.05 level. There were significant changes in the visibility of fibers and specks between different kVp values with W/Rh (p < 0.001). However, with W/Ag combination, significant differences were observed in the fibers only (p < 0.001). Among the kVp values used, 28 kV emerged as the optimal value. Comparison of images obtained with the two filter materials, led to significant differences in the visibility of fibers and specks (p < 0.008). At 32 kVp, there were significant differences in the visibility of specks only (p < 0.008). A W/Rh target-filter combination provides better image quality than that provided by W/Ag. In particular, 30 and 32 kVp X-ray beams produce higher quality images than the lower kV values. Copyright © 2017 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.

  13. Image quality improvement in cone-beam CT using the super-resolution technique.

    PubMed

    Oyama, Asuka; Kumagai, Shinobu; Arai, Norikazu; Takata, Takeshi; Saikawa, Yusuke; Shiraishi, Kenshiro; Kobayashi, Takenori; Kotoku, Jun'ichi

    2018-04-05

    This study was conducted to improve cone-beam computed tomography (CBCT) image quality using the super-resolution technique, a method of inferring a high-resolution image from a low-resolution image. This technique is used with two matrices, so-called dictionaries, constructed respectively from high-resolution and low-resolution image bases. For this study, a CBCT image, as a low-resolution image, is represented as a linear combination of atoms, the image bases in the low-resolution dictionary. The corresponding super-resolution image was inferred by multiplying the coefficients and the high-resolution dictionary atoms extracted from planning CT images. To evaluate the proposed method, we computed the root mean square error (RMSE) and structural similarity (SSIM). The resulting RMSE and SSIM between the super-resolution images and the planning CT images were, respectively, as much as 0.81 and 1.29 times better than those obtained without using the super-resolution technique. We used super-resolution technique to improve the CBCT image quality.

  14. Implications of Fuel-Efficient Vehicles on Ride Quality and Passenger Acceptance : Workshop Proceedings, Woods Hole, Massachusetts, September 6-8, 1978

    DOT National Transportation Integrated Search

    1979-08-01

    Four workshops were conducted under the auspices of the Transportation Research Board. The topics of discussion included ride quality and passenger acceptance problems associated with enhanced fuel efficiency of automobiles (Group A) and aircraft (Gr...

  15. TU-F-9A-01: Balancing Image Quality and Dose in Radiography

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Peck, D; Pasciak, A

    2014-06-15

    Emphasis is often placed on minimizing radiation dose in diagnostic imaging without a complete consideration of the effect on image quality, especially those that affect diagnostic accuracy. This session will include a patient image-based review of diagnostic quantities important to radiologists in conventional radiography, including the effects of body habitus, age, positioning, and the clinical indication of the exam. The relationships between image quality, radiation dose, and radiation risk will be discussed, specifically addressing how these factors are affected by image protocols and acquisition parameters and techniques. This session will also discuss some of the actual and perceived radiation riskmore » associated with diagnostic imaging. Regardless if the probability for radiation-induced cancer is small, the fear associated with radiation persists. Also when a risk has a benefit to an individual or to society, the risk may be justified with respect to the benefit. But how do you convey the risks and the benefits to people? This requires knowledge of how people perceive risk and how to communicate the risk and the benefit to different populations. In this presentation the sources of errors in estimating risk from radiation and some methods used to convey risks are reviewed. Learning Objectives: Understand the image quality metrics that are clinically relevant to radiologists. Understand how acquisition parameters and techniques affect image quality and radiation dose in conventional radiology. Understand the uncertainties in estimates of radiation risk from imaging exams. Learn some methods for effectively communicating radiation risk to the public.« less

  16. Assessing the image quality of pelvic MR images acquired with a flat couch for radiotherapy treatment planning.

    PubMed

    McJury, M; O'Neill, A; Lawson, M; McGrath, C; Grey, A; Page, W; O'Sullivan, J M

    2011-08-01

    To improve the integration of MRI with radiotherapy treatment planning, our department fabricated a flat couch top for our MR scanner. Setting up using this couch top meant that the patients were physically higher up in the scanner and, posteriorly, a gap was introduced between the patient and radiofrequency coil. Phantom measurements were performed to assess the quantitative impact on image quality. A phantom was set up with and without the flat couch insert in place, and measurements of image uniformity and signal to noise were made. To assess clinical impact, six patients with pelvic cancer were recruited and scanned on both couch types. The image quality of pairs of scans was assessed by two consultant radiologists. The use of the flat couch insert led to a drop in image signal to noise of approximately 14%. Uniformity in the anteroposterior direction was affected the most, with little change in right-to-left and feet-to-head directions. All six patients were successfully scanned on the flat couch, although one patient had to be positioned with their arms by their sides. The image quality scores showed no statistically significant change between scans with and without the flat couch in place. Although the quantitative performance of the coil is affected by the integration of a flat couch top, there is no discernible deterioration of diagnostic image quality, as assessed by two consultant radiologists. Although the flat couch insert moved patients higher in the bore of the scanner, all patients in the study were successfully scanned.

  17. Kilovoltage cone-beam CT: Comparative dose and image quality evaluations in partial and full-angle scan protocols

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kim, Sangroh; Yoo, Sua; Yin Fangfang

    2010-07-15

    protocols were found as {approx}14 times lower for standard head scan and 1.8 times lower for standard body scan than the old protocols, respectively. In the image quality QA tests, all the protocols except low-dose head and low-dose thorax protocols were within the tolerance in the HU verification test. The HU value for the two protocols was always higher than the nominal value. All the protocols passed the spatial linearity/resolution and HU uniformity tests. In the contrast resolution test, only high-quality head and pelvis scan protocols were within the tolerance. In addition, crescent effect was found in the partial-angle scan protocols. Conclusions: The authors found that CTDI{sub w} of the new CBCT protocols has been significantly reduced compared to the old protocols with acceptable image quality. The CTDI{sub w} values in the point dose method were close to the volume averaging method within 9%-21% for all the CBCT scan protocols. The Bakalyar's method produced more accurate dose estimation within 14%. The HU inaccuracy from low-dose head and low-dose thorax protocols can render incorrect dose results in the treatment planning system. When high soft-tissue contrast data are desired, high-quality head or pelvis scan protocol is recommended depending on the imaging area. The point dose method can be applicable to estimate CBCT dose with reasonable accuracy in the clinical environment.« less

  18. A worksite vegan nutrition program is well-accepted and improves health-related quality of life and work productivity.

    PubMed

    Katcher, Heather I; Ferdowsian, Hope R; Hoover, Valerie J; Cohen, Joshua L; Barnard, Neal D

    2010-01-01

    Vegetarian and vegan diets are effective in preventing and treating several chronic diseases. However, their acceptability outside a clinical trial setting has not been extensively studied. The aim of this study was to determine the acceptability of a worksite vegan nutrition program and its effects on health-related quality of life and work productivity. Employees of a major insurance corporation with a body mass index > or =25 kg/m(2) and/or a previous diagnosis of type 2 diabetes received either weekly group instruction on a low-fat vegan diet (n = 68) or received no diet instruction (n = 45) for 22 weeks. The vegan group reported improvements in general health (p = 0.002), physical functioning (p = 0.001), mental health (p = 0.03), vitality (p = 0.004), and overall diet satisfaction (p < 0.001) compared with the control group. The vegan group also reported a decrease in food costs (p = 0.003), and increased difficulty finding foods when eating out (p = 0.04) compared with the control group. The vegan group reported a 40-46% decrease in health-related productivity impairments at work (p = 0.03) and in regular daily activities (p = 0.004). A worksite vegan nutrition program is well-accepted and can be implemented by employers to improve the health, quality of life, and work productivity of employees.

  19. Acceptability of quality reporting and pay for performance among primary health centers in Lebanon.

    PubMed

    Saleh, Shadi S; Alameddine, Mohamad S; Natafgi, Nabil M

    2013-01-01

    Primary health care (PHC) is emphasized as the cornerstone of any health care system. Enhancing PHC performance is considered a strategy to enhance effective and equitable access to care. This study assesses the acceptability of and factors associated with quality reporting among PHC centers (PHCCs) in Lebanon. The managers of 132 Lebanese Ministry of Health PHCCs were surveyed using a cross-sectional design. Managers' willingness to report quality, participate in comparative quality assessments, and endorse pay-for-performance schemes was evaluated. Collected data were matched to the infrastructural characteristics and services database. Seventy-six percent of managers responded to the questionnaire, 93 percent of whom were willing to report clinical performance. Most expressed strong support for peer-performance comparison and pay-for-performance schemes. Willingness to report was negatively associated with the religious affiliation of centers and presence of health care facilities in the catchment area and favorably associated with use of information systems and the size of population served. The great willingness of PHCC managers to employ quality-enhancing initiatives flags a policy priority for PHC stakeholders to strengthen PHCC infrastructure and to enable reporting in an easy, standardized, and systematic way. Enhancing equity necessitates education and empowerment of managers in remote areas and those managing religiously affiliated centers.

  20. No-reference multiscale blur detection tool for content based image retrieval

    NASA Astrophysics Data System (ADS)

    Ezekiel, Soundararajan; Stocker, Russell; Harrity, Kyle; Alford, Mark; Ferris, David; Blasch, Erik; Gorniak, Mark

    2014-06-01

    In recent years, digital cameras have been widely used for image capturing. These devices are equipped in cell phones, laptops, tablets, webcams, etc. Image quality is an important component of digital image analysis. To assess image quality for these mobile products, a standard image is required as a reference image. In this case, Root Mean Square Error and Peak Signal to Noise Ratio can be used to measure the quality of the images. However, these methods are not possible if there is no reference image. In our approach, a discrete-wavelet transformation is applied to the blurred image, which decomposes into the approximate image and three detail sub-images, namely horizontal, vertical, and diagonal images. We then focus on noise-measuring the detail images and blur-measuring the approximate image to assess the image quality. We then compute noise mean and noise ratio from the detail images, and blur mean and blur ratio from the approximate image. The Multi-scale Blur Detection (MBD) metric provides both an assessment of the noise and blur content. These values are weighted based on a linear regression against full-reference y values. From these statistics, we can compare to normal useful image statistics for image quality without needing a reference image. We then test the validity of our obtained weights by R2 analysis as well as using them to estimate image quality of an image with a known quality measure. The result shows that our method provides acceptable results for images containing low to mid noise levels and blur content.

  1. Professional Acceptance Of Electronic Images In Radiologic Practice

    NASA Astrophysics Data System (ADS)

    Gitlin, Joseph N.; Curtis, David J.; Kerlin, Barbara D.; Olmsted, William W.

    1983-05-01

    During the past four years, a large number of radiographic images have been interpreted in both film and video modes in an effort to determine the utility of digital/analogue systems in general practice. With the cooperation of the Department of Defense, the MITRE Corporation, and several university-based radiology departments, the Public Health Service has participated in laboratory experiments and a teleradiology field trial to meet this objective. During the field trial, 30 radiologists participated in the interpretation of more than 4,000 diagnostic x-ray examinations that were performed at distant clinics, digitized, and transmitted to a medical center for interpretation on video monitors. As part of the evaluation, all of the participating radiologists and the attending physicians at the clinics were queried regarding the teleradiology system, particularly with respect to the diagnostic quality of the electronic images. The original films for each of the 4,000 examinations were read independently, and the findings and impressions from each mode were compared to identify discrepancies. In addition, a sample of 530 cases was reviewed and interpreted by a consensus panel to measure the accuracy of findings and impressions of both film and video readings. The sample has been retained in an automated archive for future study at the National Center of Devices and Radiological Health facilities in Rockville, Maryland. The studies include a comparison of diagnostic findings and impressions from 1024 x 1024 matrices with those obtained from the 512 x 512 format used in the field trial. The archive also provides a database for determining the effect of data compression techniques on diagnostic interpretations and establishing the utility of image processing algorithms. The paper will include an analysis of the final results of the field trial and preliminary findings from the ongoing studies using the archive of cases at the National Center for Devices and Radiological

  2. A virtual image chain for perceived image quality of medical display

    NASA Astrophysics Data System (ADS)

    Marchessoux, Cédric; Jung, Jürgen

    2006-03-01

    This paper describes a virtual image chain for medical display (project VICTOR: granted in the 5th framework program by European commission). The chain starts from raw data of an image digitizer (CR, DR) or synthetic patterns and covers image enhancement (MUSICA by Agfa) and both display possibilities, hardcopy (film on viewing box) and softcopy (monitor). Key feature of the chain is a complete image wise approach. A first prototype is implemented in an object-oriented software platform. The display chain consists of several modules. Raw images are either taken from scanners (CR-DR) or from a pattern generator, in which characteristics of DR- CR systems are introduced by their MTF and their dose-dependent Poisson noise. The image undergoes image enhancement and comes to display. For soft display, color and monochrome monitors are used in the simulation. The image is down-sampled. The non-linear response of a color monitor is taken into account by the GOG or S-curve model, whereas the Standard Gray-Scale-Display-Function (DICOM) is used for monochrome display. The MTF of the monitor is applied on the image in intensity levels. For hardcopy display, the combination of film, printer, lightbox and viewing condition is modeled. The image is up-sampled and the DICOM-GSDF or a Kanamori Look-Up-Table is applied. An anisotropic model for the MTF of the printer is applied on the image in intensity levels. The density-dependent color (XYZ) of the hardcopy film is introduced by Look-Up-tables. Finally a Human Visual System Model is applied to the intensity images (XYZ in terms of cd/m2) in order to eliminate nonvisible differences. Comparison leads to visible differences, which are quantified by higher order image quality metrics. A specific image viewer is used for the visualization of the intensity image and the visual difference maps.

  3. Does clinical pretest probability influence image quality and diagnostic accuracy in dual-source coronary CT angiography?

    PubMed

    Thomas, Christoph; Brodoefel, Harald; Tsiflikas, Ilias; Bruckner, Friederike; Reimann, Anja; Ketelsen, Dominik; Drosch, Tanja; Claussen, Claus D; Kopp, Andreas; Heuschmid, Martin; Burgstahler, Christof

    2010-02-01

    To prospectively evaluate the influence of the clinical pretest probability assessed by the Morise score onto image quality and diagnostic accuracy in coronary dual-source computed tomography angiography (DSCTA). In 61 patients, DSCTA and invasive coronary angiography were performed. Subjective image quality and accuracy for stenosis detection (>50%) of DSCTA with invasive coronary angiography as gold standard were evaluated. The influence of pretest probability onto image quality and accuracy was assessed by logistic regression and chi-square testing. Correlations of image quality and accuracy with the Morise score were determined using linear regression. Thirty-eight patients were categorized into the high, 21 into the intermediate, and 2 into the low probability group. Accuracies for the detection of significant stenoses were 0.94, 0.97, and 1.00, respectively. Logistic regressions and chi-square tests showed statistically significant correlations between Morise score and image quality (P < .0001 and P < .001) and accuracy (P = .0049 and P = .027). Linear regression revealed a cutoff Morise score for a good image quality of 16 and a cutoff for a barely diagnostic image quality beyond the upper Morise scale. Pretest probability is a weak predictor of image quality and diagnostic accuracy in coronary DSCTA. A sufficient image quality for diagnostic images can be reached with all pretest probabilities. Therefore, coronary DSCTA might be suitable also for patients with a high pretest probability. Copyright 2010 AUR. Published by Elsevier Inc. All rights reserved.

  4. Image quality assessment for teledermatology: from consumer devices to a dedicated medical device

    NASA Astrophysics Data System (ADS)

    Amouroux, Marine; Le Cunff, Sébastien; Haudrechy, Alexandre; Blondel, Walter

    2017-03-01

    Aging population as well as growing incidence of type 2 diabetes induce a growing incidence of chronic skin disorders. In the meantime, chronic shortage of dermatologists leaves some areas underserved. Remote triage and assistance to homecare nurses (known as "teledermatology") appear to be promising solutions to provide dermatological valuation in a decent time to patients wherever they live. Nowadays, teledermatology is often based on consumer devices (digital tablets, smartphones, webcams) whose photobiological and electrical safety levels do not match with medical devices' levels. The American Telemedicine Association (ATA) has published recommendations on quality standards for teledermatology. This "quick guide" does not address the issue of image quality which is critical in domestic environments where lighting is rarely reproducible. Standardized approaches of image quality would allow clinical trial comparison, calibration, manufacturing quality control and quality insurance during clinical use. Therefore, we defined several critical metrics using calibration charts (color and resolution charts) in order to assess image quality such as resolution, lighting uniformity, color repeatability and discrimination of key couples of colors. Using such metrics, we compared quality of images produced by several medical devices (handheld and video-dermoscopes) as well as by consumer devices (digital tablet and cameras) widely spread among dermatologists practice. Since diagnosis accuracy may be impaired by "low quality-images", this study highlights that, from an optical point of view, teledermatology should only be performed using medical devices. Furthermore, a dedicated medical device should probably be developed for the time follow-up of skin lesions often managed in teledermatology such as chronic wounds that require i) noncontact imaging of ii) large areas of skin surfaces, both criteria that cannot be matched using dermoscopes.

  5. Computerized symptom and quality-of-life assessment for patients with cancer part II: acceptability and usability.

    PubMed

    Mullen, Kristin H; Berry, Donna L; Zierler, Brenda K

    2004-09-01

    To determine the acceptability and usability of a computerized quality-of-life (QOL) and symptom assessment tool and the graphically displayed QOL and symptom output in an ambulatory radiation oncology clinic. Descriptive, cross-sectional. Radiation oncology clinic located in an urban university medical center. 45 patients with cancer being evaluated for radiation therapy and 10 clinicians, who submitted 12 surveys. Acceptability of the computerized assessment was measured with an online, 16-item, Likert-style survey delivered as 45 patients undergoing radiation therapy completed a 25-item QOL and symptom assessment. Usability of the graphic output was assessed with clinician completion of a four-item paper survey. Acceptability and usability of computerized patient assessment. The patient acceptability survey indicated that 70% (n = 28) liked computers and 10% (n = 4) did not. The program was easy to use for 79% (n = 26), easy to understand for 91% (n = 30), and enjoyable for 71% (n = 24). Seventy-six percent (n = 25) believed that the amount of time needed to complete the computerized survey was acceptable. Sixty-six percent (n = 21) responded that they were satisfied with the program, and none of the participants chose the very dissatisfied response. Eighty-three percent (n = 10) of the clinicians found the graphic output helpful in promoting communication with patients, 75% (n = 9) found the output report helpful in identifying appropriate areas of QOL deficits or concerns, and 83% (n = 10) indicated that the output helped guide clinical interactions with patients. The computer-based QOL and symptom assessment tool is acceptable to patients, and the graphically displayed QOL and symptom output is useful to radiation oncology nurses and physicians. Wider application of computerized patient-generated data can continue in various cancer settings and be tested for clinical and organizational outcomes.

  6. A real time quality control application for animal production by image processing.

    PubMed

    Sungur, Cemil; Özkan, Halil

    2015-11-01

    Standards of hygiene and health are of major importance in food production, and quality control has become obligatory in this field. Thanks to rapidly developing technologies, it is now possible for automatic and safe quality control of food production. For this purpose, image-processing-based quality control systems used in industrial applications are being employed to analyze the quality of food products. In this study, quality control of chicken (Gallus domesticus) eggs was achieved using a real time image-processing technique. In order to execute the quality control processes, a conveying mechanism was used. Eggs passing on a conveyor belt were continuously photographed in real time by cameras located above the belt. The images obtained were processed by various methods and techniques. Using digital instrumentation, the volume of the eggs was measured, broken/cracked eggs were separated and dirty eggs were determined. In accordance with international standards for classifying the quality of eggs, the class of separated eggs was determined through a fuzzy implication model. According to tests carried out on thousands of eggs, a quality control process with an accuracy of 98% was possible. © 2014 Society of Chemical Industry.

  7. Image quality comparison of two multifocal IOLs: influence of the pupil.

    PubMed

    García-Domene, Mari Carmen; Felipe, Adelina; Peris-Martínez, Cristina; Navea, Amparo; Artigas, Jose M; Pons, Álvaro M

    2015-04-01

    To evaluate the effect of pupil size on image quality of a sectorial multifocal intraocular lens (IOL), the Lentis Mplus (Oculentis GmbH, Berlin, Germany), and the Acri.LISA IOL (Carl Zeiss Meditec, Jena, Germany). The authors measured the MTFs of the Lentis Mplus LS-312 IOL and the Acri.LISA 366D IOL with three different sizes of pupil diameters: 3, 4, and 5 mm. The MTF was calculated from the cross-line spread function recorded with the OPAL Vector System (Image Science Ltd., Oxford, UK) by using fast Fourier-transform techniques. In distance focus, the image quality provided by the Lentis Mplus IOL was better than that of the Acri. LISA IOL with all pupil diameters. In near focus, the MTF of the Acri.LISA IOL was better with a 3-mm pupil, but poor with larger pupils. The aberration effect was equal in both IOLs in distance focus, but in near focus and with a 3-mm pupil, the Acri.LISA IOL was less affected by the aberration than the Lentis Mplus IOL. The Lentis Mplus IOL provides better distance image quality than the Acri.LISA IOL, whereas the near image quality of the Acri.LISA IOL is better with small-pupil diameter. The sectorial design makes this IOL more suitable for patients with a pupil diameter greater than 3 mm. Copyright 2015, SLACK Incorporated.

  8. A survey of quality measures for gray-scale image compression

    NASA Technical Reports Server (NTRS)

    Eskicioglu, Ahmet M.; Fisher, Paul S.

    1993-01-01

    Although a variety of techniques are available today for gray-scale image compression, a complete evaluation of these techniques cannot be made as there is no single reliable objective criterion for measuring the error in compressed images. The traditional subjective criteria are burdensome, and usually inaccurate or inconsistent. On the other hand, being the most common objective criterion, the mean square error (MSE) does not have a good correlation with the viewer's response. It is now understood that in order to have a reliable quality measure, a representative model of the complex human visual system is required. In this paper, we survey and give a classification of the criteria for the evaluation of monochrome image quality.

  9. The Effect of Image Quality, Repeated Study, and Assessment Method on Anatomy Learning

    ERIC Educational Resources Information Center

    Fenesi, Barbara; Mackinnon, Chelsea; Cheng, Lucia; Kim, Joseph A.; Wainman, Bruce C.

    2017-01-01

    The use of two-dimensional (2D) images is consistently used to prepare anatomy students for handling real specimen. This study examined whether the quality of 2D images is a critical component in anatomy learning. The visual clarity and consistency of 2D anatomical images was systematically manipulated to produce low-quality and high-quality…

  10. Data Quality Evaluation and Application Potential Analysis of TIANGONG-2 Wide-Band Imaging Spectrometer

    NASA Astrophysics Data System (ADS)

    Qin, B.; Li, L.; Li, S.

    2018-04-01

    Tiangong-2 is the first space laboratory in China, which launched in September 15, 2016. Wide-band Imaging Spectrometer is a medium resolution multispectral imager on Tiangong-2. In this paper, the authors introduced the indexes and parameters of Wideband Imaging Spectrometer, and made an objective evaluation about the data quality of Wide-band Imaging Spectrometer in radiation quality, image sharpness and information content, and compared the data quality evaluation results with that of Landsat-8. Although the data quality of Wide-band Imager Spectrometer has a certain disparity with Landsat-8 OLI data in terms of signal to noise ratio, clarity and entropy. Compared with OLI, Wide-band Imager Spectrometer has more bands, narrower bandwidth and wider swath, which make it a useful remote sensing data source in classification and identification of large and medium scale ground objects. In the future, Wide-band Imaging Spectrometer data will be widely applied in land cover classification, ecological environment assessment, marine and coastal zone monitoring, crop identification and classification, and other related areas.

  11. On the subjective acceptance during cardiovascular magnetic resonance imaging at 7.0 Tesla.

    PubMed

    Klix, Sabrina; Els, Antje; Paul, Katharina; Graessl, Andreas; Oezerdem, Celal; Weinberger, Oliver; Winter, Lukas; Thalhammer, Christof; Huelnhagen, Till; Rieger, Jan; Mehling, Heidrun; Schulz-Menger, Jeanette; Niendorf, Thoralf

    2015-01-01

    This study examines the subjective acceptance during UHF-CMR in a cohort of healthy volunteers who underwent a cardiac MR examination at 7.0T. Within a period of two-and-a-half years (January 2012 to June 2014) a total of 165 healthy volunteers (41 female, 124 male) without any known history of cardiac disease underwent UHF-CMR. For the assessment of the subjective acceptance a questionnaire was used to examine the participants experience prior, during and after the UHF-CMR examination. For this purpose, subjects were asked to respond to the questionnaire in an exit interview held immediately after the completion of the UHF-CMR examination under supervision of a study nurse to ensure accurate understanding of the questions. All questions were answered with "yes" or "no" including space for additional comments. Transient muscular contraction was documented in 12.7% of the questionnaires. Muscular contraction was reported to occur only during periods of scanning with the magnetic field gradients being rapidly switched. Dizziness during the study was reported by 12.7% of the subjects. Taste of metal was reported by 10.1% of the study population. Light flashes were reported by 3.6% of the entire cohort. 13% of the subjects reported side effects/observations which were not explicitly listed in the questionnaire but covered by the question about other side effects. No severe side effects as vomiting or syncope after scanning occurred. No increase in heart rate was observed during the UHF-CMR exam versus the baseline clinical examination. This study adds to the literature by detailing the subjective acceptance of cardiovascular magnetic resonance imaging examinations at a magnetic field strength of 7.0T. Cardiac MR examinations at 7.0T are well tolerated by healthy subjects. Broader observational and multi-center studies including patient cohorts with cardiac diseases are required to gain further insights into the subjective acceptance of UHF-CMR examinations.

  12. Improvement of the clinical use of computed radiography for mobile chest imaging: Image quality and patient dose

    NASA Astrophysics Data System (ADS)

    Rill, Lynn Neitzey

    Chest radiography is technically difficult because of the wide variation of tissue attenuations in the chest and limitations of screen-film systems. Mobile chest radiography, performed bedside on hospital inpatients, presents additional difficulties due to geometrical and equipment limitations inherent to mobile x-ray procedures and the severity of illness in patients. Computed radiography (CR) offers a new approach for mobile chest radiography by utilizing a photostimulable phosphor. Photostimulable phosphors are more efficient in absorbing lower-energy x-rays than standard intensifying screens and overcome some image quality limitations of mobile chest imaging, particularly because of the inherent latitude. This study evaluated changes in imaging parameters for CR to take advantage of differences between CR and screen-film radiography. Two chest phantoms, made of acrylic and aluminum, simulated x-ray attenuation for average-sized and large- sized adult chests. The phantoms contained regions representing the lungs, heart and subdiaphragm. Acrylic and aluminum disks (1.9 cm diameter) were positioned in the chest regions to make signal-to-noise ratio (SNR) measurements for different combinations of imaging parameters. Disk thicknesses (contrast) were determined from disk visibility. Effective dose to the phantom was also measured for technique combinations. The results indicated that using an anti-scatter grid and lowering x- ray tube potential improved the SNR significantly; however, the dose to the phantom also increased. An evaluation was performed to examine the clinical applicability of the observed improvements in SNR. Parameter adjustments that improved phantom SNRs by more than 50% resulted in perceived image quality improvements in the lung region of clinical mobile chest radiographs. Parameters that produced smaller improvements in SNR had no apparent effect on clinical image quality. Based on this study, it is recommended that a 3:1 grid be used for

  13. Study of the performance of image restoration under different wavefront aberrations

    NASA Astrophysics Data System (ADS)

    Wang, Xinqiu; Hu, Xinqi

    2016-10-01

    Image restoration is an effective way to improve the quality of images degraded by wave-front aberrations. If the wave-front aberration is too large, the performance of the image restoration will not be good. In this paper, the relationship between the performance of image restoration and the degree of wave-front aberrations is studied. A set of different wave-front aberrations is constructed by Zernike polynomials, and the corresponding PSF under white-light illumination is calculated. A set of blurred images is then obtained through convolution methods. Next we recover the images with the regularized Richardson-Lucy algorithm and use the RMS of the original image and the homologous deblurred image to evaluate the quality of restoration. Consequently, we determine the range of wave-front errors in which the recovered images are acceptable.

  14. Image Quality of Cardiac Magnetic Resonance Imaging in Patients With an Implantable Cardioverter Defibrillator System Designed for the Magnetic Resonance Imaging Environment.

    PubMed

    Schwitter, Juerg; Gold, Michael R; Al Fagih, Ahmed; Lee, Sung; Peterson, Michael; Ciuffo, Allen; Zhang, Yan; Kristiansen, Nina; Kanal, Emanuel; Sommer, Torsten

    2016-05-01

    Recently, magnetic resonance (MR)-conditional implantable cardioverter defibrillator (ICD) systems have become available. However, associated cardiac MR image (MRI) quality is unknown. The goal was to evaluate the image quality performance of various cardiac MR sequences in a multicenter trial of patients implanted with an MR-conditional ICD system. The Evera-MRI trial enrolled 275 patients in 42 centers worldwide. There were 263 patients implanted with an Evera-MRI single- or dual-chamber ICD and randomized to controls (n=88) and MRI (n=175), 156 of whom underwent a protocol-required MRI (9-12 weeks post implant). Steady-state-free-precession (SSFP) and fast-gradient-echo (FGE) sequences were acquired in short-axis and horizontal long-axis orientations. Qualitative and quantitative assessment of image quality was performed by using a 7-point scale (grades 1-3: good quality, grades 6-7: nondiagnostic) and measuring ICD- and lead-related artifact size. Good to moderate image quality (grades 1-5) was obtained in 53% and 74% of SSFP and FGE acquisitions, respectively, covering the left ventricle, and in 69% and 84%, respectively, covering the right ventricle. Odds for better image quality were greater for right ventricle versus left ventricle (odds ratio, 1.8; 95% confidence interval, 1.5-2.2; P<0.0001) and greater for FGE versus SSFP (odds ratio, 3.5; 95% confidence interval, 2.5-4.8; P<0.0001). Compared with SSFP, ICD-related artifacts on FGE were smaller (141±65 versus 75±57 mm, respectively; P<0.0001). Lead artifacts were much smaller than ICD artifacts (P<0.0001). FGE yields good to moderate quality in 74% of left ventricle and 84% of right ventricle acquisitions and performs better than SSFP in patients with an MRI-conditional ICD system. In these patients, cardiac MRI can offer diagnostic information in most cases. URL: http://www.clinicaltrials.gov. Unique identifier: NCT02117414. © 2016 American Heart Association, Inc.

  15. Emperical Tests of Acceptance Sampling Plans

    NASA Technical Reports Server (NTRS)

    White, K. Preston, Jr.; Johnson, Kenneth L.

    2012-01-01

    Acceptance sampling is a quality control procedure applied as an alternative to 100% inspection. A random sample of items is drawn from a lot to determine the fraction of items which have a required quality characteristic. Both the number of items to be inspected and the criterion for determining conformance of the lot to the requirement are given by an appropriate sampling plan with specified risks of Type I and Type II sampling errors. In this paper, we present the results of empirical tests of the accuracy of selected sampling plans reported in the literature. These plans are for measureable quality characteristics which are known have either binomial, exponential, normal, gamma, Weibull, inverse Gaussian, or Poisson distributions. In the main, results support the accepted wisdom that variables acceptance plans are superior to attributes (binomial) acceptance plans, in the sense that these provide comparable protection against risks at reduced sampling cost. For the Gaussian and Weibull plans, however, there are ranges of the shape parameters for which the required sample sizes are in fact larger than the corresponding attributes plans, dramatically so for instances of large skew. Tests further confirm that the published inverse-Gaussian (IG) plan is flawed, as reported by White and Johnson (2011).

  16. Dynamic intensity-weighted region of interest imaging for conebeam CT

    PubMed Central

    Pearson, Erik; Pan, Xiaochuan; Pelizzari, Charles

    2017-01-01

    BACKGROUND Patient dose from image guidance in radiotherapy is small compared to the treatment dose. However, the imaging beam is untargeted and deposits dose equally in tumor and healthy tissues. It is desirable to minimize imaging dose while maintaining efficacy. OBJECTIVE Image guidance typically does not require full image quality throughout the patient. Dynamic filtration of the kV beam allows local control of CT image noise for high quality around the target volume and lower quality elsewhere, with substantial dose sparing and reduced scatter fluence on the detector. METHODS The dynamic Intensity-Weighted Region of Interest (dIWROI) technique spatially varies beam intensity during acquisition with copper filter collimation. Fluence is reduced by 95% under the filters with the aperture conformed dynamically to the ROI during cone-beam CT scanning. Preprocessing to account for physical effects of the collimator before reconstruction is described. RESULTS Reconstructions show image quality comparable to a standard scan in the ROI, with higher noise and streak artifacts in the outer region but still adequate quality for patient localization. Monte Carlo modeling shows dose reduction by 10–15% in the ROI due to reduced scatter, and up to 75% outside. CONCLUSIONS The presented technique offers a method to reduce imaging dose by accepting increased image noise outside the ROI, while maintaining full image quality inside the ROI. PMID:27257875

  17. Comparison of image quality and radiation dose between split-filter dual-energy images and single-energy images in single-source abdominal CT.

    PubMed

    Euler, André; Obmann, Markus M; Szucs-Farkas, Zsolt; Mileto, Achille; Zaehringer, Caroline; Falkowski, Anna L; Winkel, David J; Marin, Daniele; Stieltjes, Bram; Krauss, Bernhard; Schindera, Sebastian T

    2018-02-19

    To compare image quality and radiation dose of abdominal split-filter dual-energy CT (SF-DECT) combined with monoenergetic imaging to single-energy CT (SECT) with automatic tube voltage selection (ATVS). Two-hundred single-source abdominal CT scans were performed as SECT with ATVS (n = 100) and SF-DECT (n = 100). SF-DECT scans were reconstructed and subdivided into composed images (SF-CI) and monoenergetic images at 55 keV (SF-MI). Objective and subjective image quality were compared among single-energy images (SEI), SF-CI and SF-MI. CNR and FOM were separately calculated for the liver (e.g. CNR liv ) and the portal vein (CNR pv ). Radiation dose was compared using size-specific dose estimate (SSDE). Results of the three groups were compared using non-parametric tests. Image noise of SF-CI was 18% lower compared to SEI and 48% lower compared to SF-MI (p < 0.001). Composed images yielded higher CNR liv over single-energy images (23.4 vs. 20.9; p < 0.001), whereas CNR pv was significantly lower (3.5 vs. 5.2; p < 0.001). Monoenergetic images overcame this inferiority in CNR pv and achieved similar results compared to single-energy images (5.1 vs. 5.2; p > 0.628). Subjective sharpness was equal between single-energy and monoenergetic images and diagnostic confidence was equal between single-energy and composed images. FOM liv was highest for SF-CI. FOM pv was equal for SEI and SF-MI (p = 0.78). SSDE was significant lower for SF-DECT compared to SECT (p < 0.022). The combined use of split-filter dual-energy CT images provides comparable objective and subjective image quality at lower radiation dose compared to single-energy CT with ATVS. • Split-filter dual-energy results in 18% lower noise compared to single-energy with ATVS. • Split-filter dual-energy results in 11% lower SSDE compared to single-energy with ATVS. • Spectral shaping of split-filter dual-energy leads to an increased dose-efficiency.

  18. A Magnetic Resonance Imaging-Conditional External Cardiac Defibrillator for Resuscitation Within the Magnetic Resonance Imaging Scanner Bore.

    PubMed

    Schmidt, Ehud J; Watkins, Ronald D; Zviman, Menekhem M; Guttman, Michael A; Wang, Wei; Halperin, Henry A

    2016-10-01

    Subjects undergoing cardiac arrest within a magnetic resonance imaging (MRI) scanner are currently removed from the bore and then from the MRI suite, before the delivery of cardiopulmonary resuscitation and defibrillation, potentially increasing the risk of mortality. This precludes many higher-risk (acute ischemic and acute stroke) patients from undergoing MRI and MRI-guided intervention. An MRI-conditional cardiac defibrillator should enable scanning with defibrillation pads attached and the generator ON, enabling application of defibrillation within the seconds of MRI after a cardiac event. An MRI-conditional external defibrillator may improve patient acceptance for MRI procedures. A commercial external defibrillator was rendered 1.5 Tesla MRI-conditional by the addition of novel radiofrequency filters between the generator and commercial disposable surface pads. The radiofrequency filters reduced emission into the MRI scanner and prevented cable/surface pad heating during imaging, while preserving all the defibrillator monitoring and delivery functions. Human volunteers were imaged using high specific absorption rate sequences to validate MRI image quality and lack of heating. Swine were electrically fibrillated (n=4) and thereafter defibrillated both outside and inside the MRI bore. MRI image quality was reduced by 0.8 or 1.6 dB, with the generator in monitoring mode and operating on battery or AC power, respectively. Commercial surface pads did not create artifacts deeper than 6 mm below the skin surface. Radiofrequency heating was within US Food and Drug Administration guidelines. Defibrillation was completely successful inside and outside the MRI bore. A prototype MRI-conditional defibrillation system successfully defibrillated in the MRI without degrading the image quality or increasing the time needed for defibrillation. It can increase patient acceptance for MRI procedures. © 2016 American Heart Association, Inc.

  19. Applications of emerging imaging techniques for meat quality and safety detection and evaluation: A review.

    PubMed

    Xiong, Zhenjie; Sun, Da-Wen; Pu, Hongbin; Gao, Wenhong; Dai, Qiong

    2017-03-04

    With improvement in people's living standards, many people nowadays pay more attention to quality and safety of meat. However, traditional methods for meat quality and safety detection and evaluation, such as manual inspection, mechanical methods, and chemical methods, are tedious, time-consuming, and destructive, which cannot meet the requirements of modern meat industry. Therefore, seeking out rapid, non-destructive, and accurate inspection techniques is important for the meat industry. In recent years, a number of novel and noninvasive imaging techniques, such as optical imaging, ultrasound imaging, tomographic imaging, thermal imaging, and odor imaging, have emerged and shown great potential in quality and safety assessment. In this paper, a detailed overview of advanced applications of these emerging imaging techniques for quality and safety assessment of different types of meat (pork, beef, lamb, chicken, and fish) is presented. In addition, advantages and disadvantages of each imaging technique are also summarized. Finally, future trends for these emerging imaging techniques are discussed, including integration of multiple imaging techniques, cost reduction, and developing powerful image-processing algorithms.

  20. Image quality comparison between single energy and dual energy CT protocols for hepatic imaging

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Yao, Yuan, E-mail: yuanyao@stanford.edu; Pelc, Nor

    Purpose: Multi-detector computed tomography (MDCT) enables volumetric scans in a single breath hold and is clinically useful for hepatic imaging. For simple tasks, conventional single energy (SE) computed tomography (CT) images acquired at the optimal tube potential are known to have better quality than dual energy (DE) blended images. However, liver imaging is complex and often requires imaging of both structures containing iodinated contrast media, where atomic number differences are the primary contrast mechanism, and other structures, where density differences are the primary contrast mechanism. Hence it is conceivable that the broad spectrum used in a dual energy acquisition maymore » be an advantage. In this work we are interested in comparing these two imaging strategies at equal-dose and more complex settings. Methods: We developed numerical anthropomorphic phantoms to mimic realistic clinical CT scans for medium size and large size patients. MDCT images based on the defined phantoms were simulated using various SE and DE protocols at pre- and post-contrast stages. For SE CT, images from 60 kVp through 140 with 10 kVp steps were considered; for DE CT, both 80/140 and 100/140 kVp scans were simulated and linearly blended at the optimal weights. To make a fair comparison, the mAs of each scan was adjusted to match the reference radiation dose (120 kVp, 200 mAs for medium size patients and 140 kVp, 400 mAs for large size patients). Contrast-to-noise ratio (CNR) of liver against other soft tissues was used to evaluate and compare the SE and DE protocols, and multiple pre- and post-contrasted liver-tissue pairs were used to define a composite CNR. To help validate the simulation results, we conducted a small clinical study. Eighty-five 120 kVp images and 81 blended 80/140 kVp images were collected and compared through both quantitative image quality analysis and an observer study. Results: In the simulation study, we found that the CNR of pre-contrast SE image

  1. Can state-of-the-art HVS-based objective image quality criteria be used for image reconstruction techniques based on ROI analysis?

    NASA Astrophysics Data System (ADS)

    Dostal, P.; Krasula, L.; Klima, M.

    2012-06-01

    Various image processing techniques in multimedia technology are optimized using visual attention feature of the human visual system. Spatial non-uniformity causes that different locations in an image are of different importance in terms of perception of the image. In other words, the perceived image quality depends mainly on the quality of important locations known as regions of interest. The performance of such techniques is measured by subjective evaluation or objective image quality criteria. Many state-of-the-art objective metrics are based on HVS properties; SSIM, MS-SSIM based on image structural information, VIF based on the information that human brain can ideally gain from the reference image or FSIM utilizing the low-level features to assign the different importance to each location in the image. But still none of these objective metrics utilize the analysis of regions of interest. We solve the question if these objective metrics can be used for effective evaluation of images reconstructed by processing techniques based on ROI analysis utilizing high-level features. In this paper authors show that the state-of-the-art objective metrics do not correlate well with subjective evaluation while the demosaicing based on ROI analysis is used for reconstruction. The ROI were computed from "ground truth" visual attention data. The algorithm combining two known demosaicing techniques on the basis of ROI location is proposed to reconstruct the ROI in fine quality while the rest of image is reconstructed with low quality. The color image reconstructed by this ROI approach was compared with selected demosaicing techniques by objective criteria and subjective testing. The qualitative comparison of the objective and subjective results indicates that the state-of-the-art objective metrics are still not suitable for evaluation image processing techniques based on ROI analysis and new criteria is demanded.

  2. Patient movement characteristics and the impact on CBCT image quality and interpretability.

    PubMed

    Spin-Neto, Rubens; Costa, Cláudio; Salgado, Daniela Mra; Zambrana, Nataly Rm; Gotfredsen, Erik; Wenzel, Ann

    2018-01-01

    To assess the impact of patient movement characteristics and metal/radiopaque materials in the field-of-view (FOV) on CBCT image quality and interpretability. 162 CBCT examinations were performed in 134 consecutive (i.e. prospective data collection) patients (age average: 27.2 years; range: 9-73). An accelerometer-gyroscope system registered patient's head position during examination. The threshold for movement definition was set at ≥0.5-mm movement distance based on accelerometer-gyroscope recording. Movement complexity was defined as uniplanar/multiplanar. Three observers scored independently: presence of stripe (i.e. streak) artefacts (absent/"enamel stripes"/"metal stripes"/"movement stripes"), overall unsharpness (absent/present) and image interpretability (interpretable/not interpretable). Kappa statistics assessed interobserver agreement. χ 2 tests analysed whether movement distance, movement complexity and metal/radiopaque material in the FOV affected image quality and image interpretability. Relevant risk factors (p ≤ 0.20) were entered into a multivariate logistic regression analysis with "not interpretable" as the outcome. Interobserver agreement for image interpretability was good (average = 0.65). Movement distance and presence of metal/radiopaque materials significantly affected image quality and interpretability. There were 22-28 cases, in which the observers stated the image was not interpretable. Small movements (i.e. <3 mm) did not significantly affect image interpretability. For movements ≥ 3 mm, the risk that a case was scored as "not interpretable" was significantly (p ≤ 0.05) increased [OR 3.2-11.3; 95% CI (0.70-65.47)]. Metal/radiopaque material was also a significant (p ≤ 0.05) risk factor (OR 3.61-5.05). Patient movement ≥3 mm and metal/radiopaque material in the FOV significantly affected CBCT image quality and interpretability.

  3. Evaluation of the Spies TM modalities image quality.

    PubMed

    Emiliani, Esteban; Talso, Michele; Baghdadi, Mohammed; Barreiro, Aaron; Orosa, Andrea; Serviàn, Pol; Gavrilov, Pavel; Proietti, Silvia; Traxer, Olivier

    2017-01-01

    The Spies™ system (Karl-Storz®) was introduced into digital ureteroscopy to improve endoscopic vision. To date, there is no data to either indicate which of the Spies modalities is better for improving diagnosis and treatment procedures, nor to compare the modalities in terms of image quality. The aim of this study was to evaluate and compare the image quality of five Spies™ modalities (SM) to the standard white light in an in-vitro model. Two standardized grids and 3 stones of different composition were recorded in white light and the 5SM (Clara, Chroma, Clara+Chroma), Spectra A and B) using 4 standardized aqueous scenarios. Twelve templates were done in order to simultaneously compare the same objective in the different modalities. Six urologists, five medical students, five urology residents, and five persons not involved with urology evaluated each video on a scale of 1 (very bad) to 5 (very good). Comparing white light to SM, subjects scored better the quality of Clara and Clara+Chroma than white light (p=0.0139 and p<0.05) and scored worse Spectra A and B (p=0.0005 and p=0.0023). When comparing Clara to the other SM, it was ranked equivalent to Clara+Chroma (p=0.67) and obtained a higher rank than Chroma, Spectra A and B (p<0.05, p=0.0001 and p=0.0001). In the multivariate analysis mean scores were higher among urologists. In all analyzed scenarios, the subjects ranked Clara and Clara+Chroma as the modalities with better image quality compared to white light. Copyright® by the International Brazilian Journal of Urology.

  4. The effect of light intensity on image quality in endoscopic ear surgery.

    PubMed

    McCallum, R; McColl, J; Iyer, A

    2018-05-16

    Endoscopic ear surgery is a rapidly developing field with many advantages. But endoscopes can reach temperatures of over 110°C at the tip, raising safety concerns. Reducing the intensity of the light source reduces temperatures produced. However, quality of images at lower light intensities has not yet been studied. We set out to study the effect of light intensity on image quality in EES. Prospective study of patients undergoing EES from April to October 2016. Consecutive images of the same operative field at 10%, 30%, 50% and 100% light intensities were taken. Eight international experts were asked to each evaluate 100 anonymised, randomised images. District General Hospital. Twenty patients. Images were evaluated on a 5-point Likert scale (1 = significantly worse than average; 5 = significantly better than average) for detail of anatomy; colour contrast; overall quality; and suitability for operating. Mean scores for photographs at 10%, 30%, 50% and 100% light intensity were 3.22 (SD 0.93), 3.15 (SD 0.84), 3.08 (SD 0.88) and 3.10 (SD 0.86), respectively. In ANOVA models for the scores on each of the scales (anatomy, colour contrast, overall quality and suitability for operating), the effects of rater and patient were highly significant (P < .0005) but light intensity was non-significant (P = .34, .32, .21, .15, respectively). Images taken during surgery by our endoscope and operative camera have no loss of quality when taken at lower light intensities. We recommend the surgeon considers use of lower light intensities in endoscopic ear surgery. © 2018 John Wiley & Sons Ltd.

  5. Evaluation of data quality, timeliness and acceptability of the tuberculosis surveillance system in Brazil's micro-regions.

    PubMed

    Silva, Gabriela Drummond Marques da; Bartholomay, Patrícia; Cruz, Oswaldo Gonçalves; Garcia, Leila Posenato

    2017-10-01

    This study aimed to evaluate quality, acceptability and timeliness of the data in the tuberculosis surveillance system in Brazilian micro-regions. An ecological cross-sectional study was carried out, after a qualitative stage for selecting indicators. All 558 Brazilian micro-regions were used as units of analysis. Data available in the National Notifiable Diseases Information System (SINAN), from 2012 to 2014, were used to calculate 14 indicators relating to four attributes: completeness, consistency, timeliness and acceptability. The study made use of cluster analysis to group micro-regions according to acceptability and timeliness. Three clusters were identified among the 473 micro-regions with optimal or regular completeness (70% to 100%) and with over five notifications. Cluster 1 (n = 109) presented mean timeliness of notification and treatment equal to 62.8% and 24.9%, respectively. Cluster 2 (n = 143) had a mean percentage of cases tested for HIV equal to 55.9%. Cluster 3 (n = 221) had the best performing tuberculosis indicators. Results suggest priority areas for improving surveillance of tuberculosis, predominantly in the central-north part of the country. They also point to the need to increase the timeliness of treatment and the percentage of cases tested for HIV.

  6. Light-leaking region segmentation of FOG fiber based on quality evaluation of infrared image

    NASA Astrophysics Data System (ADS)

    Liu, Haoting; Wang, Wei; Gao, Feng; Shan, Lianjie; Ma, Yuzhou; Ge, Wenqian

    2014-07-01

    To improve the assembly reliability of Fiber Optic Gyroscope (FOG), a light leakage detection system and method is developed. First, an agile movement control platform is designed to implement the pose control of FOG optical path component in 6 Degrees of Freedom (DOF). Second, an infrared camera is employed to capture the working state images of corresponding fibers in optical path component after the manual assembly of FOG; therefore the entire light transmission process of key sections in light-path can be recorded. Third, an image quality evaluation based region segmentation method is developed for the light leakage images. In contrast to the traditional methods, the image quality metrics, including the region contrast, the edge blur, and the image noise level, are firstly considered to distinguish the image characters of infrared image; then the robust segmentation algorithms, including graph cut and flood fill, are all developed for region segmentation according to the specific image quality. Finally, after the image segmentation of light leakage region, the typical light-leaking type, such as the point defect, the wedge defect, and the surface defect can be identified. By using the image quality based method, the applicability of our proposed system can be improved dramatically. Many experiment results have proved the validity and effectiveness of this method.

  7. TH-A-16A-01: Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Seibert, J; Imbergamo, P

    The expansion and integration of diagnostic imaging technologies such as On Board Imaging (OBI) and Cone Beam Computed Tomography (CBCT) into radiation oncology has required radiation oncology physicists to be responsible for and become familiar with assessing image quality. Unfortunately many radiation oncology physicists have had little or no training or experience in measuring and assessing image quality. Many physicists have turned to automated QA analysis software without having a fundamental understanding of image quality measures. This session will review the basic image quality measures of imaging technologies used in the radiation oncology clinic, such as low contrast resolution, highmore » contrast resolution, uniformity, noise, and contrast scale, and how to measure and assess them in a meaningful way. Additionally a discussion of the implementation of an image quality assurance program in compliance with Task Group recommendations will be presented along with the advantages and disadvantages of automated analysis methods. Learning Objectives: Review and understanding of the fundamentals of image quality. Review and understanding of the basic image quality measures of imaging modalities used in the radiation oncology clinic. Understand how to implement an image quality assurance program and to assess basic image quality measures in a meaningful way.« less

  8. 'It's good enough': Swedish general dental practitioners on reasons for accepting substandard root filling quality.

    PubMed

    Dahlström, L; Lindwall, O; Rystedt, H; Reit, C

    2018-04-01

    The concept of 'good enough' is central and necessary in the assessment of root filling quality. The aim was to explore the concept by analysing reasons and arguments for the acceptance or rejection of substandard root filling quality as reported by general dental practitioners (GDPs) in Sweden. The study was designed as a qualitative and exploratory study based on seven videotaped focus group interviews analysed by means of qualitative content analysis. Thirty-three GDPs employed in the Public Dental Health Service in Gothenburg, Sweden, participated (4-6 GDPs/interview). In all, nine predetermined questions were followed. Before each focus group, the participants received radiographs of 37 root fillings and were asked to assess the root filling quality. The three cases representing the most divergent assessments served as a basis for the discussion. The cases were presented without clinical information; the dentists would relate to the cases as being just root filled by themselves. The radiographs did not provide a sufficient basis for decisions on whether or not to accept the root filling. This study emphasized that dentists did not primarily look for these arguments in the technical details of the root filling per se, but instead, they considered selected features of the contextual situation. The GDPs constantly introduced relevant 'ad hoc considerations' to account for the decisions they made. These contextual considerations were related to aspects of pulpal and periapical disease, risks (e.g. technical complications) or to consumed resources (personal and/or economic). It was obvious that the concept of 'good enough' does not exist as a general formula ready to be applied in particular situations. Instead, it is necessarily and irremediably tied to contextual properties that emerge from case to case. © 2017 International Endodontic Journal. Published by John Wiley & Sons Ltd.

  9. Image quality characteristics of a novel colour scanning digital ophthalmoscope (SDO) compared with fundus photography.

    PubMed

    Strauss, Rupert W; Krieglstein, Tina R; Priglinger, Siegfried G; Reis, Werner; Ulbig, Michael W; Kampik, Anselm; Neubauer, Aljoscha S

    2007-11-01

    To establish a set of quality parameters for grading image quality and apply those to evaluate the fundus image quality obtained by a new scanning digital ophthalmoscope (SDO) compared with standard slide photography. On visual analogue scales a total of eight image characteristics were defined: overall quality, contrast, colour brilliance, focus (sharpness), resolution and details, noise, artefacts and validity of clinical assessment. Grading was repeated after 4 months to assess repeatability. Fundus images of 23 patients imaged digitally by SDO and by Zeiss 450FF fundus camera using Kodak film were graded side-by-side by three graders. Lens opacity was quantified with the Interzeag Lens Opacity Meter 701. For all of the eight scales of image quality, good repeatability within the graders (mean Kendall's W 0.69) was obtained after 4 months. Inter-grader agreement ranged between 0.31 and 0.66. Despite the SDO's limited nominal image resolution of 720 x 576 pixels, the Zeiss FF 450 camera performed better in only two of the subscales - noise (p = 0.001) and artefacts (p = 0.01). Lens opacities significantly influenced only the two subscales 'resolution' and 'details', which deteriorated with increasing media opacities for both imaging systems. Distinct scales to grade image characteristics of different origin were developed and validated. Overall SDO digital imaging was found to provide fundus pictures of a similarly high level of quality as expert photography on slides.

  10. The Cucurbitaceae of India: Accepted names, synonyms, geographic distribution, and information on images and DNA sequences

    PubMed Central

    Renner, Susanne S.; Pandey, Arun K.

    2013-01-01

    Abstract The most recent critical checklists of the Cucurbitaceae of India are 30 years old. Since then, botanical exploration, online availability of specimen images and taxonomic literature, and molecular-phylogenetic studies have led to modified taxon boundaries and geographic ranges. We present a checklist of the Cucurbitaceae of India that treats 400 relevant names and provides information on the collecting locations and herbaria for all types. We accept 94 species (10 of them endemic) in 31 genera. For accepted species, we provide their geographic distribution inside and outside India, links to online images of herbarium or living specimens, and information on publicly available DNA sequences to highlight gaps in the current understanding of Indian cucurbit diversity. Of the 94 species, 79% have DNA sequences in GenBank, albeit rarely from Indian material. The most species-rich genera are Trichosanthes with 22 species, Cucumis with 11 (all but two wild), Momordica with 8, and Zehneria with 5. From an evolutionary point of view, India is of special interest because it harbors a wide range of lineages, many of them relatively old and phylogenetically isolated. Phytogeographically, the north eastern and peninsular regions are richest in species, while the Jammu Kashmir and Himachal regions have few Cucurbitaceae. Our checklist probably underestimates the true diversity of Indian Cucurbitaceae, but should help focus efforts towards the least known species and regions. PMID:23717193

  11. SU-E-T-60: A Plan Quality Index in IMRT QA That Is Independent of the Acceptance Criteria

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kim, D; Kang, S; Kim, T

    2015-06-15

    Purpose: In IMRT QA, plan quality evaluation is made based on pass rate under preset acceptance criteria, mostly using gamma-values. This method is convenient but, its Result highly depends on what the acceptance criteria are and suffers from the lack of sensitivity in judging how good the plan is. In this study, we introduced a simple but effective plan quality index of IMRT QA based on dose difference only to supplement such shortcomings, and investigated its validity. Methods: The proposed index is a single value which is calculated mainly based on point-by-point comparison between planned and measured dose distributions, andmore » it becomes “1” in an ideal case. A systematic evaluation was performed with one-dimensional test dose distributions. For 3 hypothetical dose profiles, various displacements (in both dose and space) were introduced, the proposed index was calculated for each case, and the behavior of obtained indices was analyzed and compared with that of gamma evaluation. In addition, the feasibility of the index was assessed with clinical IMRT/VMAT/SBRT QA cases for different sites (prostate, head & neck, liver, lung, spine, and abdomen). Results: The proposed index showed more robust correlation with the amount of induced displacement compared to the gamma evaluation method. No matter what the acceptance criteria are (e.g., whether 3%/3mm or 2%/2mm), it was possible to clearly rank every case with the proposed index while it was difficult to do with the gamma evaluation method. Conclusion: IMRT plan quality can be evaluated quantitatively by the proposed index. It is considered that the proposed index would provide useful information for better judging the level of goodness of each plan and its Result is independent of the acceptance criteria. This work was supported by the Radiation Technology R&D program (No. 2013M2A2A7043498) and the Mid-career Researcher Program (2014R1A2A1A10050270) through the National Research Foundation of Korea funded

  12. Developing Matlab scripts for image analysis and quality assessment

    NASA Astrophysics Data System (ADS)

    Vaiopoulos, A. D.

    2011-11-01

    Image processing is a very helpful tool in many fields of modern sciences that involve digital imaging examination and interpretation. Processed images however, often need to be correlated with the original image, in order to ensure that the resulting image fulfills its purpose. Aside from the visual examination, which is mandatory, image quality indices (such as correlation coefficient, entropy and others) are very useful, when deciding which processed image is the most satisfactory. For this reason, a single program (script) was written in Matlab language, which automatically calculates eight indices by utilizing eight respective functions (independent function scripts). The program was tested in both fused hyperspectral (Hyperion-ALI) and multispectral (ALI, Landsat) imagery and proved to be efficient. Indices were found to be in agreement with visual examination and statistical observations.

  13. Importance of methodology on (99m)technetium dimercapto-succinic acid scintigraphic image quality: imaging pilot study for RIVUR (Randomized Intervention for Children With Vesicoureteral Reflux) multicenter investigation.

    PubMed

    Ziessman, Harvey A; Majd, Massoud

    2009-07-01

    We reviewed our experience with (99m)technetium dimercapto-succinic acid scintigraphy obtained during an imaging pilot study for a multicenter investigation (Randomized Intervention for Children With Vesicoureteral Reflux) of the effectiveness of daily antimicrobial prophylaxis for preventing recurrent urinary tract infection and renal scarring. We analyzed imaging methodology and its relation to diagnostic image quality. (99m)Technetium dimercapto-succinic acid imaging guidelines were provided to participating sites. High-resolution planar imaging with parallel hole or pinhole collimation was required. Two core reviewers evaluated all submitted images. Analysis included appropriate views, presence or lack of patient motion, adequate magnification, sufficient counts and diagnostic image quality. Inter-reader agreement was evaluated. We evaluated 70, (99m)technetium dimercapto-succinic acid studies from 14 institutions. Variability was noted in methodology and image quality. Correlation (r value) between dose administered and patient age was 0.780. For parallel hole collimator imaging good correlation was noted between activity administered and counts (r = 0.800). For pinhole imaging the correlation was poor (r = 0.110). A total of 10 studies (17%) were rejected for quality issues of motion, kidney overlap, inadequate magnification, inadequate counts and poor quality images. The submitting institution was informed and provided with recommendations for improving quality, and resubmission of another study was required. Only 4 studies (6%) were judged differently by the 2 reviewers, and the differences were minor. Methodology and image quality for (99m)technetium dimercapto-succinic acid scintigraphy varied more than expected between institutions. The most common reason for poor image quality was inadequate count acquisition with insufficient attention to the tradeoff between administered dose, length of image acquisition, start time of imaging and resulting image

  14. Parameter Search Algorithms for Microwave Radar-Based Breast Imaging: Focal Quality Metrics as Fitness Functions.

    PubMed

    O'Loughlin, Declan; Oliveira, Bárbara L; Elahi, Muhammad Adnan; Glavin, Martin; Jones, Edward; Popović, Milica; O'Halloran, Martin

    2017-12-06

    Inaccurate estimation of average dielectric properties can have a tangible impact on microwave radar-based breast images. Despite this, recent patient imaging studies have used a fixed estimate although this is known to vary from patient to patient. Parameter search algorithms are a promising technique for estimating the average dielectric properties from the reconstructed microwave images themselves without additional hardware. In this work, qualities of accurately reconstructed images are identified from point spread functions. As the qualities of accurately reconstructed microwave images are similar to the qualities of focused microscopic and photographic images, this work proposes the use of focal quality metrics for average dielectric property estimation. The robustness of the parameter search is evaluated using experimental dielectrically heterogeneous phantoms on the three-dimensional volumetric image. Based on a very broad initial estimate of the average dielectric properties, this paper shows how these metrics can be used as suitable fitness functions in parameter search algorithms to reconstruct clear and focused microwave radar images.

  15. Using collective expert judgements to evaluate quality measures of mass spectrometry images.

    PubMed

    Palmer, Andrew; Ovchinnikova, Ekaterina; Thuné, Mikael; Lavigne, Régis; Guével, Blandine; Dyatlov, Andrey; Vitek, Olga; Pineau, Charles; Borén, Mats; Alexandrov, Theodore

    2015-06-15

    Imaging mass spectrometry (IMS) is a maturating technique of molecular imaging. Confidence in the reproducible quality of IMS data is essential for its integration into routine use. However, the predominant method for assessing quality is visual examination, a time consuming, unstandardized and non-scalable approach. So far, the problem of assessing the quality has only been marginally addressed and existing measures do not account for the spatial information of IMS data. Importantly, no approach exists for unbiased evaluation of potential quality measures. We propose a novel approach for evaluating potential measures by creating a gold-standard set using collective expert judgements upon which we evaluated image-based measures. To produce a gold standard, we engaged 80 IMS experts, each to rate the relative quality between 52 pairs of ion images from MALDI-TOF IMS datasets of rat brain coronal sections. Experts' optional feedback on their expertise, the task and the survey showed that (i) they had diverse backgrounds and sufficient expertise, (ii) the task was properly understood, and (iii) the survey was comprehensible. A moderate inter-rater agreement was achieved with Krippendorff's alpha of 0.5. A gold-standard set of 634 pairs of images with accompanying ratings was constructed and showed a high agreement of 0.85. Eight families of potential measures with a range of parameters and statistical descriptors, giving 143 in total, were evaluated. Both signal-to-noise and spatial chaos-based measures performed highly with a correlation of 0.7 to 0.9 with the gold standard ratings. Moreover, we showed that a composite measure with the linear coefficients (trained on the gold standard with regularized least squares optimization and lasso) showed a strong linear correlation of 0.94 and an accuracy of 0.98 in predicting which image in a pair was of higher quality. The anonymized data collected from the survey and the Matlab source code for data processing can be found

  16. Using collective expert judgements to evaluate quality measures of mass spectrometry images

    PubMed Central

    Palmer, Andrew; Ovchinnikova, Ekaterina; Thuné, Mikael; Lavigne, Régis; Guével, Blandine; Dyatlov, Andrey; Vitek, Olga; Pineau, Charles; Borén, Mats; Alexandrov, Theodore

    2015-01-01

    Motivation: Imaging mass spectrometry (IMS) is a maturating technique of molecular imaging. Confidence in the reproducible quality of IMS data is essential for its integration into routine use. However, the predominant method for assessing quality is visual examination, a time consuming, unstandardized and non-scalable approach. So far, the problem of assessing the quality has only been marginally addressed and existing measures do not account for the spatial information of IMS data. Importantly, no approach exists for unbiased evaluation of potential quality measures. Results: We propose a novel approach for evaluating potential measures by creating a gold-standard set using collective expert judgements upon which we evaluated image-based measures. To produce a gold standard, we engaged 80 IMS experts, each to rate the relative quality between 52 pairs of ion images from MALDI-TOF IMS datasets of rat brain coronal sections. Experts’ optional feedback on their expertise, the task and the survey showed that (i) they had diverse backgrounds and sufficient expertise, (ii) the task was properly understood, and (iii) the survey was comprehensible. A moderate inter-rater agreement was achieved with Krippendorff’s alpha of 0.5. A gold-standard set of 634 pairs of images with accompanying ratings was constructed and showed a high agreement of 0.85. Eight families of potential measures with a range of parameters and statistical descriptors, giving 143 in total, were evaluated. Both signal-to-noise and spatial chaos-based measures performed highly with a correlation of 0.7 to 0.9 with the gold standard ratings. Moreover, we showed that a composite measure with the linear coefficients (trained on the gold standard with regularized least squares optimization and lasso) showed a strong linear correlation of 0.94 and an accuracy of 0.98 in predicting which image in a pair was of higher quality. Availability and implementation: The anonymized data collected from the survey

  17. Automatic relative RPC image model bias compensation through hierarchical image matching for improving DEM quality

    NASA Astrophysics Data System (ADS)

    Noh, Myoung-Jong; Howat, Ian M.

    2018-02-01

    The quality and efficiency of automated Digital Elevation Model (DEM) extraction from stereoscopic satellite imagery is critically dependent on the accuracy of the sensor model used for co-locating pixels between stereo-pair images. In the absence of ground control or manual tie point selection, errors in the sensor models must be compensated with increased matching search-spaces, increasing both the computation time and the likelihood of spurious matches. Here we present an algorithm for automatically determining and compensating the relative bias in Rational Polynomial Coefficients (RPCs) between stereo-pairs utilizing hierarchical, sub-pixel image matching in object space. We demonstrate the algorithm using a suite of image stereo-pairs from multiple satellites over a range stereo-photogrammetrically challenging polar terrains. Besides providing a validation of the effectiveness of the algorithm for improving DEM quality, experiments with prescribed sensor model errors yield insight into the dependence of DEM characteristics and quality on relative sensor model bias. This algorithm is included in the Surface Extraction through TIN-based Search-space Minimization (SETSM) DEM extraction software package, which is the primary software used for the U.S. National Science Foundation ArcticDEM and Reference Elevation Model of Antarctica (REMA) products.

  18. Causes of cine image quality deterioration in cardiac catheterization laboratories.

    PubMed

    Levin, D C; Dunham, L R; Stueve, R

    1983-10-01

    Deterioration of cineangiographic image quality can result from malfunctions or technical errors at a number of points along the cine imaging chain: generator and automatic brightness control, x-ray tube, x-ray beam geometry, image intensifier, optics, cine camera, cine film, film processing, and cine projector. Such malfunctions or errors can result in loss of image contrast, loss of spatial resolution, improper control of film optical density (brightness), or some combination thereof. While the electronic and photographic technology involved is complex, physicians who perform cardiac catheterization should be conversant with the problems and what can be done to solve them. Catheterization laboratory personnel have control over a number of factors that directly affect image quality, including radiation dose rate per cine frame, kilovoltage or pulse width (depending on type of automatic brightness control), cine run time, selection of small or large focal spot, proper object-intensifier distance and beam collimation, aperture of the cine camera lens, selection of cine film, processing temperature, processing immersion time, and selection of developer.

  19. The Role of Peer Influence and Perceived Quality of Teaching in Faculty Acceptance of Web-Based Learning Management Systems

    ERIC Educational Resources Information Center

    Salajan, Florin D.; Welch, Anita G.; Ray, Chris M.; Peterson, Claudette

    2015-01-01

    This study's primary investigation is the impact of "peer influence" and "perceived quality of teaching" on faculty members' usage of web-based learning management systems within the Technology Acceptance Model (TAM) framework. These factors are entered into an extended TAM as external variables impacting on the core constructs…

  20. Impact of Altering Various Image Parameters on Human Epidermal Growth Factor Receptor 2 Image Analysis Data Quality.

    PubMed

    Pantanowitz, Liron; Liu, Chi; Huang, Yue; Guo, Huazhang; Rohde, Gustavo K

    2017-01-01

    The quality of data obtained from image analysis can be directly affected by several preanalytical (e.g., staining, image acquisition), analytical (e.g., algorithm, region of interest [ROI]), and postanalytical (e.g., computer processing) variables. Whole-slide scanners generate digital images that may vary depending on the type of scanner and device settings. Our goal was to evaluate the impact of altering brightness, contrast, compression, and blurring on image analysis data quality. Slides from 55 patients with invasive breast carcinoma were digitized to include a spectrum of human epidermal growth factor receptor 2 (HER2) scores analyzed with Visiopharm (30 cases with score 0, 10 with 1+, 5 with 2+, and 10 with 3+). For all images, an ROI was selected and four parameters (brightness, contrast, JPEG2000 compression, out-of-focus blurring) then serially adjusted. HER2 scores were obtained for each altered image. HER2 scores decreased with increased illumination, higher compression ratios, and increased blurring. HER2 scores increased with greater contrast. Cases with HER2 score 0 were least affected by image adjustments. This experiment shows that variations in image brightness, contrast, compression, and blurring can have major influences on image analysis results. Such changes can result in under- or over-scoring with image algorithms. Standardization of image analysis is recommended to minimize the undesirable impact such variations may have on data output.

  1. Assessment of phantom dosimetry and image quality of i-CAT FLX cone-beam computed tomography.

    PubMed

    Ludlow, John B; Walker, Cameron

    2013-12-01

    The increasing use of cone-beam computed tomography in orthodontics has been coupled with heightened concern about the long-term risks of x-ray exposure in orthodontic populations. An industry response to this has been to offer low-exposure alternative scanning options in newer cone-beam computed tomography models. Effective doses resulting from various combinations of field of view size and field location comparing child and adult anthropomorphic phantoms with the recently introduced i-CAT FLX cone-beam computed tomography unit (Imaging Sciences, Hatfield, Pa) were measured with optical stimulated dosimetry using previously validated protocols. Scan protocols included high resolution (360° rotation, 600 image frames, 120 kV[p], 5 mA, 7.4 seconds), standard (360°, 300 frames, 120 kV[p], 5 mA, 3.7 seconds), QuickScan (180°, 160 frames, 120 kV[p], 5 mA, 2 seconds), and QuickScan+ (180°, 160 frames, 90 kV[p], 3 mA, 2 seconds). Contrast-to-noise ratio was calculated as a quantitative measure of image quality for the various exposure options using the QUART DVT phantom. Child phantom doses were on average 36% greater than adult phantom doses. QuickScan+ protocols resulted in significantly lower doses than standard protocols for the child (P = 0.0167) and adult (P = 0.0055) phantoms. The 13 × 16-cm cephalometric fields of view ranged from 11 to 85 μSv in the adult phantom and 18 to 120 μSv in the child phantom for the QuickScan+ and standard protocols, respectively. The contrast-to-noise ratio was reduced by approximately two thirds when comparing QuickScan+ with standard exposure parameters. QuickScan+ effective doses are comparable with conventional panoramic examinations. Significant dose reductions are accompanied by significant reductions in image quality. However, this trade-off might be acceptable for certain diagnostic tasks such as interim assessment of treatment results. Copyright © 2013 American Association of Orthodontists. Published by Mosby, Inc

  2. Improving best-phase image quality in cardiac CT by motion correction with MAM optimization

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rohkohl, Christopher; Bruder, Herbert; Stierstorfer, Karl

    2013-03-15

    Purpose: Research in image reconstruction for cardiac CT aims at using motion correction algorithms to improve the image quality of the coronary arteries. The key to those algorithms is motion estimation, which is currently based on 3-D/3-D registration to align the structures of interest in images acquired in multiple heart phases. The need for an extended scan data range covering several heart phases is critical in terms of radiation dose to the patient and limits the clinical potential of the method. Furthermore, literature reports only slight quality improvements of the motion corrected images when compared to the most quiet phasemore » (best-phase) that was actually used for motion estimation. In this paper a motion estimation algorithm is proposed which does not require an extended scan range but works with a short scan data interval, and which markedly improves the best-phase image quality. Methods: Motion estimation is based on the definition of motion artifact metrics (MAM) to quantify motion artifacts in a 3-D reconstructed image volume. The authors use two different MAMs, entropy, and positivity. By adjusting the motion field parameters, the MAM of the resulting motion-compensated reconstruction is optimized using a gradient descent procedure. In this way motion artifacts are minimized. For a fast and practical implementation, only analytical methods are used for motion estimation and compensation. Both the MAM-optimization and a 3-D/3-D registration-based motion estimation algorithm were investigated by means of a computer-simulated vessel with a cardiac motion profile. Image quality was evaluated using normalized cross-correlation (NCC) with the ground truth template and root-mean-square deviation (RMSD). Four coronary CT angiography patient cases were reconstructed to evaluate the clinical performance of the proposed method. Results: For the MAM-approach, the best-phase image quality could be improved for all investigated heart phases, with a

  3. Speeding up image quality improvement in random phase-free holograms using ringing artifact characteristics.

    PubMed

    Nagahama, Yuki; Shimobaba, Tomoyoshi; Kakue, Takashi; Masuda, Nobuyuki; Ito, Tomoyoshi

    2017-05-01

    A holographic projector utilizes holography techniques. However, there are several barriers to realizing holographic projections. One is deterioration of hologram image quality caused by speckle noise and ringing artifacts. The combination of the random phase-free method and the Gerchberg-Saxton (GS) algorithm has improved the image quality of holograms. However, the GS algorithm requires significant computation time. We propose faster methods for image quality improvement of random phase-free holograms using the characteristics of ringing artifacts.

  4. Image quality and diagnostic performance of free-breathing diffusion-weighted imaging for hepatocellular carcinoma.

    PubMed

    Takayama, Yukihisa; Nishie, Akihiro; Asayama, Yoshiki; Ishigami, Kousei; Kakihara, Daisuke; Ushijima, Yasuhiro; Fujita, Nobuhiro; Shirabe, Ken; Takemura, Atsushi; Honda, Hiroshi

    2017-05-18

    To retrospectively evaluate the diagnostic performance of free-breathing diffusion-weighted imaging (FB-DWI) with modified imaging parameter settings for detecting hepatocellular carcinomas (HCCs). Fifty-one patients at risk for HCC were scanned with both FB-DWI and respiratory-triggered DWI with the navigator echo respiratory-triggering technique (RT-DWI). Qualitatively, the sharpness of the liver contour, the image noise and the chemical shift artifacts on each DWI with b -values of 1000 s/mm 2 were independently evaluated by three radiologists using 4-point scoring. We compared the image quality scores of each observer between the two DWI methods, using the Wilcoxon signed-rank test. Quantitatively, we compared the signal-to-noise ratios (SNRs) of the liver parenchyma and lesion-to-nonlesion contrast-to-noise ratios (CNRs) after measuring the signal intensity on each DWI with a b-factor of 1000 s/mm 2 . The average SNRs and CNRs between the two DWI methods were compared by the paired t-test. The detectability of HCC on each DWI was also analyzed by three radiologists. The detectability provided by the two DWI methods was compared using McNemar's test. For all observers, the averaged image quality scores of FB-DWI were: Sharpness of the liver contour [observer (Obs)-1, 3.08 ± 0.81; Obs-2, 2.98 ± 0.73; Obs-3, 3.54 ± 0.75], those of the distortion (Obs-1, 2.94 ± 0.50; Obs-2, 2.71 ± 0.70; Obs-3, 3.27 ± 0.53), and the chemical shift artifacts (Obs-1, 3.38 ± 0.60; Obs-2, 3.15 ± 1.07; Obs-3, 3.21 ± 0.85). The averaged image quality scores of RT-DWI were: Sharpness of the liver contour (Obs-1, 2.33 ± 0.65; Obs-2, 2.37 ± 0.74; Obs-3, 2.75 ± 0.81), distortion (Obs-1, 2.81 ± 0.56; Obs-2, 2.25 ± 0.74; Obs-3, 2.96 ± 0.71), and the chemical shift artifacts (Obs-1, 2.92 ± 0.59; Obs-2, 2.21 ± 0.85; Obs-3, 2.77 ± 1.08). All image quality scores of FB-DWI were significantly higher than those of RT-DWI ( P < 0.05). The average SNR of the normal liver parenchyma

  5. Image quality and diagnostic performance of free-breathing diffusion-weighted imaging for hepatocellular carcinoma

    PubMed Central

    Takayama, Yukihisa; Nishie, Akihiro; Asayama, Yoshiki; Ishigami, Kousei; Kakihara, Daisuke; Ushijima, Yasuhiro; Fujita, Nobuhiro; Shirabe, Ken; Takemura, Atsushi; Honda, Hiroshi

    2017-01-01

    AIM To retrospectively evaluate the diagnostic performance of free-breathing diffusion-weighted imaging (FB-DWI) with modified imaging parameter settings for detecting hepatocellular carcinomas (HCCs). METHODS Fifty-one patients at risk for HCC were scanned with both FB-DWI and respiratory-triggered DWI with the navigator echo respiratory-triggering technique (RT-DWI). Qualitatively, the sharpness of the liver contour, the image noise and the chemical shift artifacts on each DWI with b-values of 1000 s/mm2 were independently evaluated by three radiologists using 4-point scoring. We compared the image quality scores of each observer between the two DWI methods, using the Wilcoxon signed-rank test. Quantitatively, we compared the signal-to-noise ratios (SNRs) of the liver parenchyma and lesion-to-nonlesion contrast-to-noise ratios (CNRs) after measuring the signal intensity on each DWI with a b-factor of 1000 s/mm2. The average SNRs and CNRs between the two DWI methods were compared by the paired t-test. The detectability of HCC on each DWI was also analyzed by three radiologists. The detectability provided by the two DWI methods was compared using McNemar’s test. RESULTS For all observers, the averaged image quality scores of FB-DWI were: Sharpness of the liver contour [observer (Obs)-1, 3.08 ± 0.81; Obs-2, 2.98 ± 0.73; Obs-3, 3.54 ± 0.75], those of the distortion (Obs-1, 2.94 ± 0.50; Obs-2, 2.71 ± 0.70; Obs-3, 3.27 ± 0.53), and the chemical shift artifacts (Obs-1, 3.38 ± 0.60; Obs-2, 3.15 ± 1.07; Obs-3, 3.21 ± 0.85). The averaged image quality scores of RT-DWI were: Sharpness of the liver contour (Obs-1, 2.33 ± 0.65; Obs-2, 2.37 ± 0.74; Obs-3, 2.75 ± 0.81), distortion (Obs-1, 2.81 ± 0.56; Obs-2, 2.25 ± 0.74; Obs-3, 2.96 ± 0.71), and the chemical shift artifacts (Obs-1, 2.92 ± 0.59; Obs-2, 2.21 ± 0.85; Obs-3, 2.77 ± 1.08). All image quality scores of FB-DWI were significantly higher than those of RT-DWI (P < 0.05). The average SNR of the normal

  6. 48 CFR 446.370 - Inspection and acceptance.

    Code of Federal Regulations, 2010 CFR

    2010-10-01

    ... 48 Federal Acquisition Regulations System 4 2010-10-01 2010-10-01 false Inspection and acceptance. 446.370 Section 446.370 Federal Acquisition Regulations System DEPARTMENT OF AGRICULTURE CONTRACT MANAGEMENT QUALITY ASSURANCE Contract Clauses 446.370 Inspection and acceptance. The Contracting Officer...

  7. TU-B-19A-01: Image Registration II: TG132-Quality Assurance for Image Registration

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Brock, K; Mutic, S

    2014-06-15

    AAPM Task Group 132 was charged with a review of the current approaches and solutions for image registration in radiotherapy and to provide recommendations for quality assurance and quality control of these clinical processes. As the results of image registration are always used as the input of another process for planning or delivery, it is important for the user to understand and document the uncertainty associate with the algorithm in general and the Result of a specific registration. The recommendations of this task group, which at the time of abstract submission are currently being reviewed by the AAPM, include themore » following components. The user should understand the basic image registration techniques and methods of visualizing image fusion. The disclosure of basic components of the image registration by commercial vendors is critical in this respect. The physicists should perform end-to-end tests of imaging, registration, and planning/treatment systems if image registration is performed on a stand-alone system. A comprehensive commissioning process should be performed and documented by the physicist prior to clinical use of the system. As documentation is important to the safe implementation of this process, a request and report system should be integrated into the clinical workflow. Finally, a patient specific QA practice should be established for efficient evaluation of image registration results. The implementation of these recommendations will be described and illustrated during this educational session. Learning Objectives: Highlight the importance of understanding the image registration techniques used in their clinic. Describe the end-to-end tests needed for stand-alone registration systems. Illustrate a comprehensive commissioning program using both phantom data and clinical images. Describe a request and report system to ensure communication and documentation. Demonstrate an clinically-efficient patient QA practice for efficient evaluation

  8. Toward a perceptual video-quality metric

    NASA Astrophysics Data System (ADS)

    Watson, Andrew B.

    1998-07-01

    The advent of widespread distribution of digital video creates a need for automated methods for evaluating the visual quality of digital video. This is particularly so since most digital video is compressed using lossy methods, which involve the controlled introduction of potentially visible artifacts. Compounding the problem is the bursty nature of digital video, which requires adaptive bit allocation based on visual quality metrics, and the economic need to reduce bit-rate to the lowest level that yields acceptable quality. In previous work, we have developed visual quality metrics for evaluating, controlling,a nd optimizing the quality of compressed still images. These metrics incorporate simplified models of human visual sensitivity to spatial and chromatic visual signals. Here I describe a new video quality metric that is an extension of these still image metrics into the time domain. Like the still image metrics, it is based on the Discrete Cosine Transform. An effort has been made to minimize the amount of memory and computation required by the metric, in order that might be applied in the widest range of applications. To calibrate the basic sensitivity of this metric to spatial and temporal signals we have made measurements of visual thresholds for temporally varying samples of DCT quantization noise.

  9. Quality and consumer acceptability of salt and phosphate enhanced goat loin from goats fed varying levels of pine bark.

    PubMed

    Leick, C M; Broadway, P R; Solaiman, S; Behrends, J M

    2012-03-01

    Goat loins (n=22) were evaluated to test effects of 0, 15, and 30% dietary pine bark (PB) and salt, water, and phosphate enhancement on shelf-life, shear force (WBSF) and consumer acceptability. No interactions existed between PB and enhancement. Dietary PB did not affect objective color, but enhancement increased a* and b* values (P<0.05). Thiobarbituric acid reactive substances (TBARS) increased from d 1 to d 5 of storage (P<0.0001), but were not affected by PB or enhancement. The WBSF for 30% PB was less than that of 0% PB (P=0.0199), and enhancement decreased WBSF (P=0.0010). Texture, flavor, and overall acceptability were greater (P<0.05) for 15 and 30% PB compared to 0% PB. Enhanced loin samples had greater appearance, aroma, texture, flavor, and overall acceptability scores (P<0.05). Results indicated that enhancement improved tenderness and consumer acceptability of goat loin, and PB had minimal impact on goat loin quality. Copyright © 2011 Elsevier Ltd. All rights reserved.

  10. Understanding the mediating effects of relationship quality on technology acceptance: an empirical study of e-appointment system.

    PubMed

    Chen, Shih-Chih; Liu, Shih-Chi; Li, Shing-Han; Yen, David C

    2013-12-01

    This study extends the Technology Acceptance Model (TAM) by incorporating relationship quality as a mediator to construct a comprehensive framework for understanding the influence on continuance intention in the hospital e-appointment system. A survey of 334 Taiwanese citizens who were contacted via phone or the Internet and Structural Equation Modeling (SEM) is used for path analysis and hypothesis tests. The study shows that perceived ease of use (PEOU) and perceived usefulness (PU) have significant influence on continuance intention through the mediation of relationship quality, consisting of satisfaction and trust. The direct impact of relationship quality on continuance intention is also significant. The analytical results reveal that the relationship between the hospital, patients and e-appointment users can be improved via enhancing the continued usage of e-appointment. This paper also proposes a general model to synthesize the essence of PEOU, PU, and relationship quality for explaining users' continuous intention of e-appointment.

  11. Multispectral imaging for biometrics

    NASA Astrophysics Data System (ADS)

    Rowe, Robert K.; Corcoran, Stephen P.; Nixon, Kristin A.; Ostrom, Robert E.

    2005-03-01

    Automated identification systems based on fingerprint images are subject to two significant types of error: an incorrect decision about the identity of a person due to a poor quality fingerprint image and incorrectly accepting a fingerprint image generated from an artificial sample or altered finger. This paper discusses the use of multispectral sensing as a means to collect additional information about a finger that significantly augments the information collected using a conventional fingerprint imager based on total internal reflectance. In the context of this paper, "multispectral sensing" is used broadly to denote a collection of images taken under different polarization conditions and illumination configurations, as well as using multiple wavelengths. Background information is provided on conventional fingerprint imaging. A multispectral imager for fingerprint imaging is then described and a means to combine the two imaging systems into a single unit is discussed. Results from an early-stage prototype of such a system are shown.

  12. Image Quality Characteristics of Handheld Display Devices for Medical Imaging

    PubMed Central

    Yamazaki, Asumi; Liu, Peter; Cheng, Wei-Chung; Badano, Aldo

    2013-01-01

    Handheld devices such as mobile phones and tablet computers have become widespread with thousands of available software applications. Recently, handhelds are being proposed as part of medical imaging solutions, especially in emergency medicine, where immediate consultation is required. However, handheld devices differ significantly from medical workstation displays in terms of display characteristics. Moreover, the characteristics vary significantly among device types. We investigate the image quality characteristics of various handheld devices with respect to luminance response, spatial resolution, spatial noise, and reflectance. We show that the luminance characteristics of the handheld displays are different from those of workstation displays complying with grayscale standard target response suggesting that luminance calibration might be needed. Our results also demonstrate that the spatial characteristics of handhelds can surpass those of medical workstation displays particularly for recent generation devices. While a 5 mega-pixel monochrome workstation display has horizontal and vertical modulation transfer factors of 0.52 and 0.47 at the Nyquist frequency, the handheld displays released after 2011 can have values higher than 0.63 at the respective Nyquist frequencies. The noise power spectra for workstation displays are higher than 1.2×10−5 mm2 at 1 mm−1, while handheld displays have values lower than 3.7×10−6 mm2. Reflectance measurements on some of the handheld displays are consistent with measurements for workstation displays with, in some cases, low specular and diffuse reflectance coefficients. The variability of the characterization results among devices due to the different technological features indicates that image quality varies greatly among handheld display devices. PMID:24236113

  13. IEEE International Symposium on Biomedical Imaging.

    PubMed

    2017-01-01

    The IEEE International Symposium on Biomedical Imaging (ISBI) is a scientific conference dedicated to mathematical, algorithmic, and computational aspects of biological and biomedical imaging, across all scales of observation. It fosters knowledge transfer among different imaging communities and contributes to an integrative approach to biomedical imaging. ISBI is a joint initiative from the IEEE Signal Processing Society (SPS) and the IEEE Engineering in Medicine and Biology Society (EMBS). The 2018 meeting will include tutorials, and a scientific program composed of plenary talks, invited special sessions, challenges, as well as oral and poster presentations of peer-reviewed papers. High-quality papers are requested containing original contributions to the topics of interest including image formation and reconstruction, computational and statistical image processing and analysis, dynamic imaging, visualization, image quality assessment, and physical, biological, and statistical modeling. Accepted 4-page regular papers will be published in the symposium proceedings published by IEEE and included in IEEE Xplore. To encourage attendance by a broader audience of imaging scientists and offer additional presentation opportunities, ISBI 2018 will continue to have a second track featuring posters selected from 1-page abstract submissions without subsequent archival publication.

  14. Gap Acceptance During Lane Changes by Large-Truck Drivers-An Image-Based Analysis.

    PubMed

    Nobukawa, Kazutoshi; Bao, Shan; LeBlanc, David J; Zhao, Ding; Peng, Huei; Pan, Christopher S

    2016-03-01

    This paper presents an analysis of rearward gap acceptance characteristics of drivers of large trucks in highway lane change scenarios. The range between the vehicles was inferred from camera images using the estimated lane width obtained from the lane tracking camera as the reference. Six-hundred lane change events were acquired from a large-scale naturalistic driving data set. The kinematic variables from the image-based gap analysis were filtered by the weighted linear least squares in order to extrapolate them at the lane change time. In addition, the time-to-collision and required deceleration were computed, and potential safety threshold values are provided. The resulting range and range rate distributions showed directional discrepancies, i.e., in left lane changes, large trucks are often slower than other vehicles in the target lane, whereas they are usually faster in right lane changes. Video observations have confirmed that major motivations for changing lanes are different depending on the direction of move, i.e., moving to the left (faster) lane occurs due to a slower vehicle ahead or a merging vehicle on the right-hand side, whereas right lane changes are frequently made to return to the original lane after passing.

  15. Testing the quality of images for permanent magnet desktop MRI systems using specially designed phantoms.

    PubMed

    Qiu, Jianfeng; Wang, Guozhu; Min, Jiao; Wang, Xiaoyan; Wang, Pengcheng

    2013-12-21

    Our aim was to measure the performance of desktop magnetic resonance imaging (MRI) systems using specially designed phantoms, by testing imaging parameters and analysing the imaging quality. We designed multifunction phantoms with diameters of 18 and 60 mm for desktop MRI scanners in accordance with the American Association of Physicists in Medicine (AAPM) report no. 28. We scanned the phantoms with three permanent magnet 0.5 T desktop MRI systems, measured the MRI image parameters, and analysed imaging quality by comparing the data with the AAPM criteria and Chinese national standards. Image parameters included: resonance frequency, high contrast spatial resolution, low contrast object detectability, slice thickness, geometrical distortion, signal-to-noise ratio (SNR), and image uniformity. The image parameters of three desktop MRI machines could be measured using our specially designed phantoms, and most parameters were in line with MRI quality control criterion, including: resonance frequency, high contrast spatial resolution, low contrast object detectability, slice thickness, geometrical distortion, image uniformity and slice position accuracy. However, SNR was significantly lower than in some references. The imaging test and quality control are necessary for desktop MRI systems, and should be performed with the applicable phantom and corresponding standards.

  16. [Body image and quality of life in women with polycystic ovary syndrome].

    PubMed

    Micskei, Orsolya; Deli, Tamás; Jakab, Attila; Bugán, Antal

    2014-07-06

    Polycystic ovary syndrome is the most common endocrine disease in women. Psychic consequences are significant; patients have difficulties to be pregnant and the disease disturbs the quality of life. Due to complications associated with polycystic ovary syndrome, studies on psychological state and disease perception of patients appear to be important. The aim of the authors was to assess how the disease influences quality of life and changes in body image. The authors used the Illness Intrusiveness Ratings Scale, Health-Related-Quality-of-Life Questionnaire for Women with Polycystic Ovary Syndrome and their own body image questionnaire. 121 women took part in the study. There was a negative correlation between illness intrusiveness and quality of life (r = -0.353) and a positive correlation between body image and quality of life (r = 0.614). A significant difference was observed in body satisfaction between patients who had visible body changes and those who had metabolic disturbances (p<0.05). Overweight patients who lost weight with lifestyle changes had significantly better quality of life (p<0.05). These results appear to be convincing in that in addition to gynecological care, psychotherapy could be offered to patients with polycystic ovary syndrome.

  17. Psychological impact and acceptability of magnetic resonance imaging and X-ray mammography: the MARIBS Study

    PubMed Central

    Hutton, J; Walker, L G; Gilbert, F J; Evans, D G; Eeles, R; Kwan-Lim, G E; Thompson, D; Pointon, L J; Sharp, D M; Leach, M O

    2011-01-01

    Background: As part of the Magnetic Resonance Imaging for Breast Screening (MARIBS), Study women with a family history of breast cancer were assessed psychologically to determine the relative psychological impact and acceptability of annual screening using magnetic resonance imaging (MRI) and conventional X-ray mammography (XRM). Methods: Women were assessed psychologically at baseline (4 weeks before MRI and XRM), immediately before, and immediately after, both MRI and XRM, and at follow-up (6 weeks after the scans). Results: Overall, both procedures were found to be acceptable with high levels of satisfaction (MRI, 96.3% and XRM, 97.7% NS) and low levels of psychological morbidity throughout, particularly at 6-week follow-up. Low levels of self-reported distress were reported for both procedures (MRI, 13.5% and XRM, 7.8%), although MRI was more distressing (P=0.005). Similarly, higher anticipatory anxiety was reported before MRI than before XRM (P=0.003). Relative to XRM, MRI-related distress was more likely to persist at 6 weeks after the scans in the form of intrusive MRI-related thoughts (P=0.006) and total MRI-related distress (P=0.014). More women stated that they intended to return for XRM (96.3%) than for MRI (88% P<0.0005). These effects were most marked for the first year of screening, although they were also statistically significant in subsequent years. Conclusion: Given the proven benefits of MRI in screening for breast cancer in this population, these data point to the urgent need to provide timely information and support to women undergoing MRI. PMID:21326245

  18. Psychological impact and acceptability of magnetic resonance imaging and X-ray mammography: the MARIBS Study.

    PubMed

    Hutton, J; Walker, L G; Gilbert, F J; Evans, D G; Eeles, R; Kwan-Lim, G E; Thompson, D; Pointon, L J; Sharp, D M; Leach, M O

    2011-02-15

    As part of the Magnetic Resonance Imaging for Breast Screening (MARIBS), Study women with a family history of breast cancer were assessed psychologically to determine the relative psychological impact and acceptability of annual screening using magnetic resonance imaging (MRI) and conventional X-ray mammography (XRM). Women were assessed psychologically at baseline (4 weeks before MRI and XRM), immediately before, and immediately after, both MRI and XRM, and at follow-up (6 weeks after the scans). Overall, both procedures were found to be acceptable with high levels of satisfaction (MRI, 96.3% and XRM, 97.7%; NS) and low levels of psychological morbidity throughout, particularly at 6-week follow-up. Low levels of self-reported distress were reported for both procedures (MRI, 13.5% and XRM, 7.8%), although MRI was more distressing (P=0.005). Similarly, higher anticipatory anxiety was reported before MRI than before XRM (P=0.003). Relative to XRM, MRI-related distress was more likely to persist at 6 weeks after the scans in the form of intrusive MRI-related thoughts (P=0.006) and total MRI-related distress (P=0.014). More women stated that they intended to return for XRM (96.3%) than for MRI (88%; P<0.0005). These effects were most marked for the first year of screening, although they were also statistically significant in subsequent years. Given the proven benefits of MRI in screening for breast cancer in this population, these data point to the urgent need to provide timely information and support to women undergoing MRI.

  19. Imaging industry expectations for compressed sensing in MRI

    NASA Astrophysics Data System (ADS)

    King, Kevin F.; Kanwischer, Adriana; Peters, Rob

    2015-09-01

    Compressed sensing requires compressible data, incoherent acquisition and a nonlinear reconstruction algorithm to force creation of a compressible image consistent with the acquired data. MRI images are compressible using various transforms (commonly total variation or wavelets). Incoherent acquisition of MRI data by appropriate selection of pseudo-random or non-Cartesian locations in k-space is straightforward. Increasingly, commercial scanners are sold with enough computing power to enable iterative reconstruction in reasonable times. Therefore integration of compressed sensing into commercial MRI products and clinical practice is beginning. MRI frequently requires the tradeoff of spatial resolution, temporal resolution and volume of spatial coverage to obtain reasonable scan times. Compressed sensing improves scan efficiency and reduces the need for this tradeoff. Benefits to the user will include shorter scans, greater patient comfort, better image quality, more contrast types per patient slot, the enabling of previously impractical applications, and higher throughput. Challenges to vendors include deciding which applications to prioritize, guaranteeing diagnostic image quality, maintaining acceptable usability and workflow, and acquisition and reconstruction algorithm details. Application choice depends on which customer needs the vendor wants to address. The changing healthcare environment is putting cost and productivity pressure on healthcare providers. The improved scan efficiency of compressed sensing can help alleviate some of this pressure. Image quality is strongly influenced by image compressibility and acceleration factor, which must be appropriately limited. Usability and workflow concerns include reconstruction time and user interface friendliness and response. Reconstruction times are limited to about one minute for acceptable workflow. The user interface should be designed to optimize workflow and minimize additional customer training. Algorithm

  20. Dual-energy CT in patients with abdominal malignant lymphoma: impact of noise-optimised virtual monoenergetic imaging on objective and subjective image quality.

    PubMed

    Lenga, L; Czwikla, R; Wichmann, J L; Leithner, D; Albrecht, M H; D'Angelo, T; Arendt, C T; Booz, C; Hammerstingl, R; Vogl, T J; Martin, S S

    2018-06-05

    To investigate the impact of noise-optimised virtual monoenergetic imaging (VMI+) reconstructions on quantitative and qualitative image parameters in patients with malignant lymphoma at dual-energy computed tomography (DECT) examinations of the abdomen. Thirty-five consecutive patients (mean age, 53.8±18.6 years; range, 21-82 years) with histologically proven malignant lymphoma of the abdomen were included retrospectively. Images were post-processed with standard linear blending (M_0.6), traditional VMI, and VMI+ technique at energy levels ranging from 40 to 100 keV in 10 keV increments. Signal-to-noise (SNR) and contrast-to-noise ratios (CNR) were objectively measured in lymphoma lesions. Image quality, lesion delineation, and image noise were rated subjectively by three blinded observers using five-point Likert scales. Quantitative image quality parameters peaked at 40-keV VMI+ (SNR, 15.77±7.74; CNR, 18.27±8.04) with significant differences compared to standard linearly blended M_0.6 (SNR, 7.96±3.26; CNR, 13.55±3.47) and all traditional VMI series (p<0.001). Qualitative image quality assessment revealed significantly superior ratings for image quality at 60-keV VMI+ (median, 5) in comparison with all other image series (p<0.001). Assessment of lesion delineation showed the highest rating scores for 40-keV VMI+ series (median, 5), while lowest subjective image noise was found for 100-keV VMI+ reconstructions (median, 5). Low-keV VMI+ reconstructions led to improved image quality and lesion delineation of malignant lymphoma lesions compared to standard image reconstruction and traditional VMI at abdominal DECT examinations. Copyright © 2018 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.