Science.gov

Sample records for gafchromic ebt film

  1. Calibrating page sized Gafchromic EBT3 films

    SciTech Connect

    Crijns, W.; Maes, F.; Heide, U. A. van der; Van den Heuvel, F.

    2013-01-15

    Purpose: The purpose is the development of a novel calibration method for dosimetry with Gafchromic EBT3 films. The method should be applicable for pretreatment verification of volumetric modulated arc, and intensity modulated radiotherapy. Because the exposed area on film can be large for such treatments, lateral scan errors must be taken into account. The correction for the lateral scan effect is obtained from the calibration data itself. Methods: In this work, the film measurements were modeled using their relative scan values (Transmittance, T). Inside the transmittance domain a linear combination and a parabolic lateral scan correction described the observed transmittance values. The linear combination model, combined a monomer transmittance state (T{sub 0}) and a polymer transmittance state (T{sub {infinity}}) of the film. The dose domain was associated with the observed effects in the transmittance domain through a rational calibration function. On the calibration film only simple static fields were applied and page sized films were used for calibration and measurements (treatment verification). Four different calibration setups were considered and compared with respect to dose estimation accuracy. The first (I) used a calibration table from 32 regions of interest (ROIs) spread on 4 calibration films, the second (II) used 16 ROIs spread on 2 calibration films, the third (III), and fourth (IV) used 8 ROIs spread on a single calibration film. The calibration tables of the setups I, II, and IV contained eight dose levels delivered to different positions on the films, while for setup III only four dose levels were applied. Validation was performed by irradiating film strips with known doses at two different time points over the course of a week. Accuracy of the dose response and the lateral effect correction was estimated using the dose difference and the root mean squared error (RMSE), respectively. Results: A calibration based on two films was the optimal

  2. Technical Note: Homogeneity of Gafchromic EBT2 film

    SciTech Connect

    Hartmann, Bernadette; Martisikova, Maria; Jaekel, Oliver

    2010-04-15

    Purpose: The self-developing Gafchromic EBT film is a radiochromic film, widely used for relative photon dosimetry. Recently, the manufacturer has replaced the well-investigated EBT film by the new Gafchromic EBT2 film. It has the same sensitive component and, in addition, it contains a yellow marker dye in order to protect the film against ambient light exposure and to serve as a base for corrections of small differences in film response. Furthermore, the configuration of the film layers as well as the binder material have been changed in comparison to the EBT film. When investigating the properties of EBT2 film, all characteristics were found to be similar to those of EBT film, except for the film response homogeneity. Thus, in this article special focus was put on examining the homogeneity of EBT2 film. Methods: A scan protocol established for EBT film and published previously was used. The uniformity of the film coloration was investigated for unirradiated and irradiated EBT2 film sheets. The dose response of EBT2 film was measured and the influence of film inhomogeneities on dose determination was evaluated. Results: Inhomogeneities in pixel values of up to {+-}3.7% within one film were detected. The relative inhomogeneities were found to be approximately independent of the dose. Nonuniformities of the film response lead to uncertainties in dose determination of {+-}8.7% at 1 Gy. When using net optical densities for dose calibration, uncertainties in dose determination amount to more than {+-}6%. Conclusions: EBT2 films from the lot investigated in this study show response inhomogeneities, which lead to uncertainties in dose determination exceeding the commonly accepted tolerance levels. It is important to test further EBT2 lots regarding homogeneity before using the film in clinical routine.

  3. Solar ultraviolet radiation response of EBT2 Gafchromic, radiochromic film.

    PubMed

    Butson, Ethan T; Yu, Peter K N; Butson, Martin J

    2013-11-01

    Measurement of solar ultraviolet (UV) radiation is an important aspect of dosimetry for the improved knowledge of UV exposure and its associated health related issues. EBT2 Gafchromic film has been designed by its manufacturers as an improved tool for ionizing radiation dosimetry. The film is stated as exhibiting a significant reduction in UV response. However, results have shown that when exposed to UV from the 'bottom side' i.e. from the thick laminate side, the film exhibits a sensitivity to solar UV radiation which is both measurable and accurate for UV dosimetry. Films were irradiated in this position to known solar UV exposures and results are quantified showing a reproducibility of measurement to within ±7% (1 SD) when compared to calibrated UV meters. With an exposure of 20 J cm(-2) broad spectrum solar UV, the films net OD change was found to be 0.248 OD ± 0.021 OD when analysing the results using the red channel region of an Epson V700 desktop scanner. This was compared to 0.0294 OD ± 0.0053 OD change with exposure to the same UV exposure from the top side. This means that solar UV dosimetry can be performed using EBT2 Gafchromic film utilizing the underside of the film for dosimetry. The main advantages of this film type for measurement of UV exposure is the visible colour change and thus easy analysis using a desktop scanner as well as its uniformity in response and its robust physical strength for use in outside exposure situations. PMID:24113466

  4. Evaluation of Gafchromic EBT-XD film, with comparison to EBT3 film, and application in high dose radiotherapy verification.

    PubMed

    Palmer, Antony L; Dimitriadis, Alexis; Nisbet, Andrew; Clark, Catharine H

    2015-11-21

    There is renewed interest in film dosimetry for the verification of dose delivery of complex treatments, particularly small fields, compared to treatment planning system calculations. A new radiochromic film, Gafchromic EBT-XD, is available for high-dose treatment verification and we present the first published evaluation of its use. We evaluate the new film for MV photon dosimetry, including calibration curves, performance with single- and triple-channel dosimetry, and comparison to existing EBT3 film. In the verification of a typical 25 Gy stereotactic radiotherapy (SRS) treatment, compared to TPS planned dose distribution, excellent agreement was seen with EBT-XD using triple-channel dosimetry, in isodose overlay, maximum 1.0 mm difference over 200-2400 cGy, and gamma evaluation, mean passing rate 97% at 3% locally-normalised, 1.5 mm criteria. In comparison to EBT3, EBT-XD gave improved evaluation results for the SRS-plan, had improved calibration curve gradients at high doses, and had reduced lateral scanner effect. The dimensions of the two films are identical. The optical density of EBT-XD is lower than EBT3 for the same dose. The effective atomic number for both may be considered water-equivalent in MV radiotherapy. We have validated the use of EBT-XD for high-dose, small-field radiotherapy, for routine QC and a forthcoming multi-centre SRS dosimetry intercomparison. PMID:26512917

  5. Evaluation of Gafchromic EBT-XD film, with comparison to EBT3 film, and application in high dose radiotherapy verification

    NASA Astrophysics Data System (ADS)

    Palmer, Antony L.; Dimitriadis, Alexis; Nisbet, Andrew; Clark, Catharine H.

    2015-11-01

    There is renewed interest in film dosimetry for the verification of dose delivery of complex treatments, particularly small fields, compared to treatment planning system calculations. A new radiochromic film, Gafchromic EBT-XD, is available for high-dose treatment verification and we present the first published evaluation of its use. We evaluate the new film for MV photon dosimetry, including calibration curves, performance with single- and triple-channel dosimetry, and comparison to existing EBT3 film. In the verification of a typical 25 Gy stereotactic radiotherapy (SRS) treatment, compared to TPS planned dose distribution, excellent agreement was seen with EBT-XD using triple-channel dosimetry, in isodose overlay, maximum 1.0 mm difference over 200-2400 cGy, and gamma evaluation, mean passing rate 97% at 3% locally-normalised, 1.5 mm criteria. In comparison to EBT3, EBT-XD gave improved evaluation results for the SRS-plan, had improved calibration curve gradients at high doses, and had reduced lateral scanner effect. The dimensions of the two films are identical. The optical density of EBT-XD is lower than EBT3 for the same dose. The effective atomic number for both may be considered water-equivalent in MV radiotherapy. We have validated the use of EBT-XD for high-dose, small-field radiotherapy, for routine QC and a forthcoming multi-centre SRS dosimetry intercomparison.

  6. Evaluation of gafchromic EBT film for intensity modulated radiation therapy dose distribution verification

    PubMed Central

    Sankar, A.; Kurup, P. G. Goplakrishna; Murali, V.; Ayyangar, Komanduri M.; Nehru, R. Mothilal; Velmurugan, J.

    2006-01-01

    This work was undertaken with the intention of investigating the possibility of clinical use of commercially available self-developing radiochromic filmGafchromic EBT film – for IMRT dose verification. The dose response curves were generated for the films using VXR-16 film scanner. The results obtained with EBT films were compared with the results of Kodak EDR2 films. It was found that the EBT film has a linear response between the dose ranges of 0 and 600 cGy. The dose-related characteristics of the EBT film, like post-irradiation color growth with time, film uniformity and effect of scanning orientation, were studied. There is up to 8.6% increase in the color density between 2 and 40 h after irradiation. There was a considerable variation, up to 8.5%, in the film uniformity over its sensitive region. The quantitative difference between calculated and measured dose distributions was analyzed using Gamma index with the tolerance of 3% dose difference and 3 mm distance agreement. EDR2 films showed good and consistent results with the calculated dose distribution, whereas the results obtained using EBT were inconsistent. The variation in the film uniformity limits the use of EBT film for conventional large field IMRT verification. For IMRT of smaller field size (4.5 × 4.5 cm), the results obtained with EBT were comparable with results of EDR2 films. PMID:21206669

  7. Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT GAFCHROMIC[reg] film

    SciTech Connect

    Lynch, Bart D.; Kozelka, Jakub; Ranade, Manisha K.; Li, Jonathan G.; Simon, William E.; Dempsey, James F.

    2006-12-15

    In this study, we present three significant artifacts that have the potential to negatively impact the accuracy and precision of film dosimetry measurements made using GAFCHROMIC[reg] EBT radiochromic film when read out with CCD flatbed scanners. Films were scanned using three commonly employed instruments: a Macbeth TD932 spot densitometer, an Epson Expression 1680 CCD array scanner, and a Microtek ScanMaker i900 CCD array scanner. For the two scanners we assessed the variation in optical density (OD) of GAFCHROMIC EBT film with scanning bed position, angular rotation of the film with respect to the scan line direction, and temperature inside the scanner due to repeated scanning. Scanning uniform radiochromic films demonstrated a distinct bowing effect in profiles in the direction of the CCD array with a nonuniformity of up to 17%. Profiles along a direction orthogonal to the CCD array demonstrated a 7% variation. A strong angular dependence was found in measurements made with the flatbed scanners; the effect could not be reproduced with the spot densitometer. An IMRT quality assurance film was scanned twice rotating the film 90 deg. between the scans. For films scanned on the Epson scanner, up to 12% variation was observed in unirradiated EBT films rotated between 0 deg. and 90 deg. , which decreased to approximately 8% for EBT films irradiated to 300 cGy. Variations of up to 80% were observed for films scanned with the Microtek scanner. The scanners were found to significantly increase the film temperature with repeated scanning. Film temperature between 18 and 33 deg. C caused OD changes of approximately 7%. Considering these effects, we recommend adherence to a strict scanning protocol that includes: maintaining the orientation of films scanned on flatbed scanners, limiting scanning to the central portion of the scanner bed, and limiting the number of consecutive scans to minimize changes in OD caused by film heating.

  8. SU-E-T-76: Comparing Homogeneity Between Gafchromic Film EBT2 and EBT3

    SciTech Connect

    Mizuno, H; Sumida, I; Ogawa, K; Tanaka, A

    2014-06-01

    Purpose: We found out that homogeneity of EBT2 was different among lot numbers in previous study. Variation in local homogeneity of EBT3 among several lot numbers has not been reported. In this study, we investigated film homogeneity of Gafcrhomic EBT3 films compared with EBT2 films. Methods: All sheets from five lots were cut into 12 pieces to investigate film homogeneity, and were irradiated at 0.5, 2, and 3 Gy. To investigate intra- and inter-sheet uniformity, five sheets from five lots were exposed to 2 Gy: intra-sheet uniformity was evaluated by the coefficient of variation of homogeneity for all pieces of a single sheet, and inter-sheet uniformity was evaluated by the coefficient of variation of homogeneity among the same piece numbers in the five sheets. To investigate the difference of ADC value in various doses, a single sheet from each of five lots was irradiated at 0.5 Gy and 3 Gy in addition to 2 Gy. A scan resolution of 72 dots per inch (dpi) and color depth of 48-bit RGB were used. Films were analyzed by the inhouse software; Average of ADC value in center ROI and profile X and Y axis were measured. Results and Conclusion: Intra-sheet uniformity of non-irradiated EBT2 films were ranged from 0.1% to 0.4%, however that of irradiated EBT2 films were ranged from 0.2% to 1.5%. On the other hand, intra-sheet uniformity of irradiated and non-irradiated EBT3 films were from 0.2% to 0.6%. Inter-sheet uniformity of all films were less than 0.5%. It was interesting point that homogeneity of EBT3 between no-irradiated and irradiated films were similar value, whereas EBT2 had dose dependence of homogeneity in ADC value evaluation. These results suggested that EBT3 homogeneity was corrected by this feature.

  9. Thermophysical investigation of Gafchromic EBT2 films using photoacoustic spectroscopy

    NASA Astrophysics Data System (ADS)

    Aydarous, A.; Abdallah, S.; Towairqi, M. Al

    2014-07-01

    The thermophysical properties of EBT2 films exposed to different doses of x-ray were investigated. The doses ranged from 2 to 818 cGy. The films were irradiated by a Varian linear accelerator using a 6 MV photon beam. The thermal conductivity (k) was obtained by measuring the thermal diffusivity (α) and thermal effusivity (e) using the photoacoustic (PA) technique. The α, e, and k values clearly indicated their dependence on the dose from 0 to 818 cGy. The results demonstrate that the PA technique can detect variations in the thermal diffusivity at doses as low as approximately 3 cGy. The thermal conductivity for the film exposed to 818 cGy of radiation increased by a factor of approximately 3.70 compared to the non-exposed film. The PA spectroscopic technique displayed good reproducibility, with a relative standard deviation of less than 5%.

  10. Technical Note: Initial characterization of the new EBT-XD Gafchromic film

    SciTech Connect

    Grams, Michael P. Gustafson, Jon M.; Long, Kenneth M.; Fong de los Santos, Luis E.

    2015-10-15

    Purpose: To assess the dosimetric accuracy and energy dependence of the new EBT-eXtended Dose (XD) Gafchromic film and to compare the lateral response artifact (LRA) between EBT-XD and EBT3 film. Methods: EBT3 and EBT-XD calibration curves were created by exposing films to known doses from 0 to 3000 cGy using a 6 MV beam. To assess the accuracy and dynamic range of EBT-XD, a 60° enhanced dynamic wedge (EDW) was used to deliver a dose range of approximately 200–2900 cGy. Comparison to treatment planning system (TPS) calculation was made using a gamma analysis with 2%/2 mm passing criteria. To assess and compare the LRA between EBT3 and EBT-XD, 21 × 21 cm{sup 2} open fields delivered doses of 1000, 2000, and 3000 cGy to both types of film. Films were placed at the center of the scanner, and ratios of measured to TPS predicted doses were calculated at 50 and 80 mm lateral from the scanner center in order to quantitatively assess the LRA. To evaluate the energy dependence of EBT-XD film, seven known doses ranging from 400 to 3000 cGy were delivered using both 6 and 18 MV beams and the resulting optical densities (ODs) compared. Results: The gamma passing rate was 99.1% for the 6 MV EDW delivery. EBT-XD film exhibited minimal LRA (<1%) up to 3000 cGy. In contrast, EBT3 demonstrated an under-response of 11.3% and 22.7% at lateral positions of 50 and 80 mm, respectively, for the 3000 cGy exposure. Differences between ODs of the EBT-XD films exposed to known doses from 6 to 18 MV beams were <0.8% suggesting minimal energy dependence throughout this energy range. Conclusions: The LRA of EBT-XD is greatly reduced when compared to EBT3. This in combination with its accuracy from 0 to 3000 cGy and minimal energy dependence from 6 to 18 MV makes EBT-XD film well suited for dosimetric measurements in high dose SRS/SBRT applications.

  11. Quality assurance of alpha-particle dosimetry using peeled-off Gafchromic EBTfilm

    NASA Astrophysics Data System (ADS)

    Ng, C. Y. P.; Chun, S. L.; Yu, K. N.

    2016-08-01

    A novel alpha-particle dosimetry technique using Gafchromic EBT3 film has recently been proposed for calibrating the activity of alpha-emitting radiopharmaceuticals. In the present paper, we outlined four measures which could further help assure the quality of the method. First, we suggested an alternative method in fabricating the peeled-off EBT3 film. Films with a chosen size were cut from the original films and all the edges were sealed with silicone. These were immersed into deionized water for 19 d and the polyester covers of the EBT3 films could then be easily peeled off. The active layers in these peeled-off EBT3 films remained intact, and these films could be prepared reproducibly with ease. Second, we proposed a check on the integrity of the peeled-off film by comparing the responses of the pristine and peeled-off EBT3 films to the same X-ray irradiation. Third, we highlighted the importance of scanning directions of the films. The "landscape" and "portrait" scanning directions were defined as the scanning directions perpendicular and parallel to the long edge of the original EBT3 films, respectively. Our results showed that the responses were different for different scanning directions. As such, the same scanning direction should be used every time. Finally, we cautioned the need to confirm the uniformity of the alpha-particle source used for calibration. Radiochromic films are well known for their capability of providing two-dimensional dosimetric information. As such, EBT3 films could also be conveniently used to check the uniformity of the alpha-particle source.

  12. Gafchromic EBT2 film dosimetry in reflection mode with a novel plan-based calibration method

    SciTech Connect

    Mendez, I.; Hartman, V.; Hudej, R.; Strojnik, A.; Casar, B.

    2013-01-15

    Purpose:A dosimetric system formed by Gafchromic EBT2 radiochromic film and Epson Expression 10000XL flatbed scanner was commissioned for dosimetry. In this paper, several open questions concerning the commissioning of radiochromic films for dosimetry were addressed: (a) is it possible to employ this dosimetric system in reflection mode; (b) if so, can the methods used in transmission mode also be used in reflection mode; (c) is it possible to obtain accurate absolute dose measurements with Gafchromic EBT2 films; (d) which calibration method should be followed; (e) which calibration models should be used; and (f) does three-color channel dosimetry offer a significant improvement over single channel dosimetry. The purpose of this paper is to help clarify these questions. Methods: In this study, films were scanned in reflection mode, the effect of surrounding film was evaluated and the feasibility of EBT2 film dosimetry in reflection mode was studied. EBT2's response homogeneity has been reported to lead to excessive dose uncertainties. To overcome this problem, a new plan-based calibration method was implemented. Plan-based calibration can use every pixel and each of the three color channels of the scanned film to obtain the parameters of the calibration model. A model selection analysis was conducted to select lateral correction and sensitometric curve models. The commonly used calibration with fragments was compared with red-channel plan-based calibration and with three-channel plan-based calibration. Results: No effect of surrounding film was found in this study. The film response inhomogeneity in EBT2 films was found to be important not only due to differences in the fog but also due to differences in sensitivity. The best results for lateral corrections were obtained using absolute corrections independent of the dose. With respect to the sensitometric curves, an empirical polynomial fit of order 4 was found to obtain results equivalent to a gamma

  13. Gafchromic EBT3 film dosimetry in electron beams - energy dependence and improved film read-out.

    PubMed

    Sipilä, Petri; Ojala, Jarkko; Kaijaluoto, Sampsa; Jokelainen, Ilkka; Kosunen, Antti

    2016-01-01

    For megavoltage photon radiation, the fundamental dosimetry characteristics of Gafchromic EBT3 film were determined in 60Co gamma ray beam with addition of experimental and Monte Carlo (MC)-simulated energy dependence of the film for 6 MV photon beam and 6 MeV, 9 MeV, 12 MeV, and 16 MeV electron beams in water phantom. For the film read-out, two phase correction of scanner sensitivity was applied: a matrix correction for scanning area and dose-dependent correction by iterative procedure. With these corrections, the uniformity of response can be improved to be within ± 50 pixel values (PVs). To improve the read-out accuracy, a procedure with flipped film orientations was established. With the method, scanner uniformity can be improved further and dust particles, scratches and/or dirt on scan-ner glass can be detected and eliminated. Responses from red and green channels were averaged for read-out, which decreased the effect of noise present in values from separate channels. Since the signal level with the blue channel is considerably lower than with other channels, the signal variation due to different perturbation effects increases the noise level so that the blue channel is not recommended to be used for dose determination. However, the blue channel can be used for the detection of emulsion thickness variations for film quality evaluations with unexposed films. With electron beams ranging from 6 MeV to 16 MeV and at reference measurement conditions in water, the energy dependence of the EBT3 film is uniform within 0.5%, with uncertainties close to 1.6% (k = 2). Including 6 MV photon beam and the electron beams mentioned, the energy dependence is within 1.1%. No notable differences were found between the experimental and MC-simulated responses, indicating negligible change in intrinsic energy dependence of the EBT3 film for 6 MV photon beam and 6 MeV-16 MeV electron beams. Based on the dosimetric characteristics of the EBT3 film, the read-out procedure established

  14. Evaluation of GAFCHROMIC registered EBT film for CyberKnife registered dosimetry

    SciTech Connect

    Wilcox, Ellen E.; Daskalov, George M.

    2007-06-15

    External beam therapy (EBT) GAFCHROMIC registered film is evaluated for dosimetry and characterization of the CyberKnife registered radiation beams. Percentage depth doses, lateral beam profiles, and output factors are measured in solid water using EBT GAFCHROMIC registered film (International Specialty Products, Wayne, NJ) for the 6 MV radiation beams of diameter 5 to 60 mm produced by the CyberKnife registered (Accuray, Sunnyvale, CA). The data are compared to those measured with the PTW 60008 diode and the Wellhofer CC01 ion chamber in water. For the small radiation field sizes used in stereotactic radiosurgery, lateral electronic disequilibrium and steep dose gradients exist in a large portion of these fields, requiring the use of high-resolution measurement techniques. For small beams, the detector size approaches the dimensions of the beam and adversely affects measurement accuracy in regions where the gradient varies across the detector. When film is the detector, the scanning system is usually the resolution-limiting component. Radiographic films based upon silver halide (AgH) emulsions are widely used for relative dosimetry of external radiation treatment beams in the megavoltage energy range, because of their good spatial resolution and capability to provide integrated dosimetry over two dimensions. Film dosimetry, however, has drawbacks due to its steep energy dependence at low photon energies as well as film processor and densitometer artifacts. EBT radiochromic film, introduced in 2004 specifically for IMRT dosimetry, may be a detector of choice for the characterization of small radiosurgical beams, because of its near-tissue equivalence, radiation beam energy independence, high spatial resolution, and self developing properties. For radiation beam sizes greater than 10 mm, the film measurements were identical to those of the diode and ion chamber. For the smaller beam diameters of 7.5 and 5 mm, however, there were differences in the data measured with

  15. SU-E-T-322: The Evaluation of the Gafchromic EBT3 Film in Low Dose 6 MV X-Ray Beams with Different Scanning Modes

    SciTech Connect

    Lee, H; Sung, J; Yoon, M; Kim, D; Chung, W

    2014-06-01

    Purpose: We have evaluated the response of the Gafchromic EBT3 film in low dose for 6 MV x-ray beams with two scanning modes, the reflection scanning mode and the transmission scanning mode. Methods: We irradiated the Gafcromic EBT3 film using a 60 degree enhanced dynamic wedge (EDW) with 6 MV x-ray beams from Clinac iX Linear accelerator (Varian Medical Systems, Palo Alto, CA). The irradiated Gafchromic EBT3 film was scanned with different scanning modes, the reflection scanning mode and the transmission scanning mode. The scanned Gafchromic EBT3 film was analyzed with MATLAB. Results: When 7.2 cGy was irradiated to the Gafchromic EBT3 film, the uncertainty was 0.54 cGy with reflection scanning mode and was 0.88 cGy with transmission scanning mode. When 24 cGy was irradiated to the Gafchromic EBT3 film, the uncertainty was similar to the case of 7.2 cGy irradiation showing 0.51 cGy of uncertainty with reflection scanning mode and 0.87 cGy of uncertainty with transmission scanning mode. The result suggests that the reflection mode should be used in Gafchromic EBT3 film for low irradiation. Conclusion: The result suggests that the reflection mode should be used in Gafchromic EBT3 film for low irradiation.

  16. Measurement of high energy x-ray beam penumbra with Gafchromic trade mark sign EBT radiochromic film

    SciTech Connect

    Cheung Tsang; Butson, Martin J.; Yu, Peter K. N.

    2006-08-15

    High energy x-ray beam penumbra are measured using Gafchromic trade mark sign EBT film. Gafchromic trade mark sign EBT, due to its limited energy dependence and high spatial resolution provide a high level of accuracy for dose assessment in penumbral regions. The spatial resolution of film detector systems is normally limited by the scanning resolution of the densitometer. Penumbral widths (80%/20%) measured at D{sub max} were found to be 2.8, 3.0, 3.2, and 3.4 mm ({+-}0.2 mm) using 5, 10, 20, and 30 cm square field sizes, respectively, for a 6 MV linear accelerator produced x-ray beam. This is compared to 3.2 mm{+-}0.2 mm (Kodak EDR2) and 3.6 mm{+-}0.2 mm (Kodak X-Omat V) at 10 cmx10 cm measured using radiographic film. Using a zero volume extrapolation technique for ionization chamber measurements, the 10 cmx10 cm field penumbra at D{sub max} was measured to be 3.1 mm, a close match to Gafchromic trade mark sign EBT results. Penumbral measurements can also be made at other depths, including the surface, as the film does not suffer significantly from dosimetric variations caused by changing x-ray energy spectra. Gafchromic trade mark sign EBT film provides an adequate measure of penumbral dose for high energy x-ray beams.

  17. EBT GAFCHROMIC{sup TM} film dosimetry in compensator-based intensity modulated radiation therapy

    SciTech Connect

    Vaezzadeh, Seyedali; Allahverdi, Mahmoud; Nedaie, Hasan A.; Ay, Mohammadreza; Shirazi, Alireza; Yarahmadi, Mehran

    2013-07-01

    The electron benefit transfer (EBT) GAFCHROMIC films possess a number of features making them appropriate for high-quality dosimetry in intensity-modulated radiation therapy (IMRT). Compensators to deliver IMRT are known to change the beam-energy spectrum as well as to produce scattered photons and to contaminate electrons; therefore, the accuracy and validity of EBT-film dosimetry in compensator-based IMRT should be investigated. Percentage-depth doses and lateral-beam profiles were measured using EBT films in perpendicular orientation with respect to 6 and 18 MV photon beam energies for: (1) different thicknesses of cerrobend slab (open, 1.0, 2.0, 4.0, and 6.0 cm), field sizes (5×5, 10×10, and 20×20 cm{sup 2}), and measurement depths (D{sub max}, 5.0 and 10.0 cm); and (2) step-wedged compensator in a solid phantom. To verify results, same measurements were implemented using a 0.125 cm{sup 3} ionization chamber in a water phantom and also in Monte Carlo simulations using the Monte Carlo N-particle radiation transport computer code. The mean energy of photons was increased due to beam hardening in comparison with open fields at both 6 and 18 MV energies. For a 20×20 cm{sup 2} field size of a 6 MV photon beam and a 6.0 cm thick block, the surface dose decreased by about 12% and percentage-depth doses increased up to 3% at 30.0 cm depth, due to the beam-hardening effect induced by the block. In contrast, at 18 MV, the surface dose increased by about 8% and depth dose reduced by 3% at 30.0 cm depth. The penumbral widths (80% to 20%) increase with block thickness, field size, and beam energy. The EBT film results were in good agreement with the ionization chamber dose profiles and Monte Carlo N-particle radiation transport computer code simulation behind the step-wedged compensator. Also, there was a good agreement between the EBT-film and the treatment-planning results on the anthropomorphic phantom. The EBT films can be accurately used as a 2D dosimeter for dose

  18. Optical imaging analysis of microscopic radiation dose gradients in Gafchromic EBT film using a digital microscope.

    PubMed

    Keller, Brian M; Peressotti, Chris; Pignol, Jean-Philippe

    2008-08-01

    By providing superior localization and immobilization, stereotactic radiosurgery (SRS) is capable of delivering millimeter spheres of dose to intracranial targets with submillimeter precision. Several authors have proposed new SRS solutions to dramatically reduce beam penumbra to hundreds of microns. These solutions require new quality assurance methods capable of penumbra measurement at the micron scale. This article examines the capability of a digital microscope, with translation stage and associated software, to resolve dose gradients in Gafchromic EBT film at this level. To produce very steep penumbra, films were irradiated in phantom beneath pinhole collimators using lower energy x rays (100 kVp, 300 kVp, and Iridium-192) and minimal geometric penumbra contribution. For film analysis, a method was developed which improved the signal to noise ratio by finding the center of the irradiation spot, generating several radial dose profiles and averaging these to obtain the final off-axis dose profile. Optical density was converted to dose using a calibration curve. The experimentally determined off-axis dose profiles were compared with MCNP Monte Carlo simulations which replicated the irradiation geometry and served to validate our measured data. The measured 80%-20% penumbral widths were 46 microm +/- 26 microm (100 kVp, 2 mm field size), 69 microm +/- m 27 microm (300 kVp, 2 mm field size), and 241 microm +/-31 microm (Ir-192, 1 mm field size). These penumbral widths agreed with Monte Carlo simulations within experimental uncertainty. Our findings suggest that reading Gafchromic EBT films using a digital microscope with translation stage is suitable for the quality assurance of very sharp penumbra able to resolve gradients to within at least 30 microm. PMID:18777933

  19. Optical imaging analysis of microscopic radiation dose gradients in Gafchromic EBT film using a digital microscope

    SciTech Connect

    Keller, Brian M.; Peressotti, Chris; Pignol, Jean-Philippe

    2008-08-15

    By providing superior localization and immobilization, stereotactic radiosurgery (SRS) is capable of delivering millimeter spheres of dose to intracranial targets with submillimeter precision. Several authors have proposed new SRS solutions to dramatically reduce beam penumbra to hundreds of microns. These solutions require new quality assurance methods capable of penumbra measurement at the micron scale. This article examines the capability of a digital microscope, with translation stage and associated software, to resolve dose gradients in Gafchromic EBT film at this level. To produce very steep penumbra, films were irradiated in phantom beneath pinhole collimators using lower energy x rays (100 kVp, 300 kVp, and Iridium-192) and minimal geometric penumbra contribution. For film analysis, a method was developed which improved the signal to noise ratio by finding the center of the irradiation spot, generating several radial dose profiles and averaging these to obtain the final off-axis dose profile. Optical density was converted to dose using a calibration curve. The experimentally determined off-axis dose profiles were compared with MCNP Monte Carlo simulations which replicated the irradiation geometry and served to validate our measured data. The measured 80%-20% penumbral widths were 46 {mu}m{+-}26 {mu}m (100 kVp, 2 mm field size), 69 {mu}m{+-}27 {mu}m (300 kVp, 2 mm field size), and 241 {mu}m{+-}31 {mu}m (Ir-192, 1 mm field size). These penumbral widths agreed with Monte Carlo simulations within experimental uncertainty. Our findings suggest that reading Gafchromic EBT films using a digital microscope with translation stage is suitable for the quality assurance of very sharp penumbra able to resolve gradients to within at least 30 {mu}m.

  20. Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification

    SciTech Connect

    Sankar, A. . E-mail: asankar_phy@yahoo.co.in; Ayyangar, Komanduri M.; Nehru, R. Mothilal; Gopalakrishna Kurup, P.G.; Murali, V.; Enke, Charles A.; Velmurugan, J.

    2006-01-01

    The quantitative dose validation of intensity-modulated radiation therapy (IMRT) plans require 2-dimensional (2D) high-resolution dosimetry systems with uniform response over its sensitive region. The present work deals with clinical use of commercially available self-developing Radio Chromic Film, Gafchromic EBT film, for IMRT dose verification. Dose response curves were generated for the films using a VXR-16 film scanner. The results obtained with EBT films were compared with the results of Kodak extended dose range 2 (EDR2) films. The EBT film had a linear response between the dose range of 0 to 600 cGy. The dose-related characteristics of the EBT film, such as post irradiation color growth with time, film uniformity, and effect of scanning orientation, were studied. There was up to 8.6% increase in the color density between 2 to 40 hours after irradiation. There was a considerable variation, up to 8.5%, in the film uniformity over its sensitive region. The quantitative differences between calculated and measured dose distributions were analyzed using DTA and Gamma index with the tolerance of 3% dose difference and 3-mm distance agreement. The EDR2 films showed consistent results with the calculated dose distributions, whereas the results obtained using EBT were inconsistent. The variation in the film uniformity limits the use of EBT film for conventional large-field IMRT verification. For IMRT of smaller field sizes (4.5 x 4.5 cm), the results obtained with EBT were comparable with results of EDR2 films.

  1. Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance

    NASA Astrophysics Data System (ADS)

    Fuss, Martina; Sturtewagen, Eva; DeWagter, Carlos; Georg, Dietmar

    2007-07-01

    The suitability of radiochromic EBT film was studied for high-precision clinical quality assurance (QA) by identifying the dose response for a wide range of irradiation parameters typically modified in highly-conformal treatment techniques. In addition, uncertainties associated with varying irradiation conditions were determined. EBT can be used for dose assessment of absorbed dose levels as well as relative dosimetry when compared to absolute absorbed dose calibrated using ionization chamber results. For comparison, a silver halide film (Kodak EDR-2) representing the current standard in film dosimetry was included. As an initial step a measurement protocol yielding accurate and precise results was established for a flatbed transparency scanner (Epson Expression 1680 Pro) that was utilized as a film reading instrument. The light transmission measured by the scanner was found to depend on the position of the film on the scanner plate. For three film pieces irradiated with doses of 0 Gy, ~1 Gy and ~7 Gy, the pixel values measured in portrait or landscape mode differed by 4.7%, 6.2% and 10.0%, respectively. A study of 200 film pieces revealed an excellent sheet-to-sheet uniformity. On a long time scale, the optical development of irradiated EBT film consisted of a slow but steady increase of absorbance which was not observed to cease during 4 months. Sensitometric curves of EBT films obtained under reference conditions (SSD = 95 cm, FS = 5 × 5 cm2, d = 5 cm) for 6, 10 and 25 MV photon beams did not show any energy dependence. The average separation between all curves was only 0.7%. The variation of the depth d (range 2-25 cm) in the phantom did not affect the dose response of EBT film. Also the influence of the radiation field size (range 3 × 3-40 × 40 cm2) on the sensitometric curve was not significant. For EDR-2 films maximum differences between the calibration curves reached 7-8% for X6MV and X25MV. Radiochromic EBT film, in combination with a flatbed scanner

  2. Evaluation of the Gafchromic{sup Registered-Sign} EBT2 film for the dosimetry of radiosurgical beams

    SciTech Connect

    Larraga-Gutierrez, Jose M.; Garcia-Hernandez, Diana; Garcia-Garduno, Olivia A.; Galvan de la Cruz, Olga O.; Ballesteros-Zebadua, Paola; Esparza-Moreno, Karina P.

    2012-10-15

    Purpose: Radiosurgery uses small fields and high-radiation doses to treat intra- and extracranial lesions in a single session. The lack of a lateral electronic equilibrium and the presence of high-dose gradients in these fields are challenges for adequate measurements. The availability of radiation detectors with the high spatial resolution required is restricted to only a few. Stereotactic diodes and EBT radiochromic films have been demonstrated to be good detectors for small-beam dosimetry. Because the stereotactic diode is the standard measurement for the dosimetry of radiosurgical beams, the goal of this work was to perform measurements with the radiochromic film Gafchromic{sup Registered-Sign} EBT2 and compare its results with a stereotactic diode. Methods: Total scatter factors, tissue maximum, and off-axis ratios from a 6 MV small photon beams were measured using EBT2 radiochromic film in a water phantom. The film-measured data were evaluated by comparing it with the data measured with a stereotactic field diode (IBA-Dosimetry). Results: The film and diode measurements had excellent agreement. The differences between the detectors were less than or equal to 2.0% for the tissue maximum and the off-axis ratios. However, for the total scatter factors, there were significant differences, up to 4.9% (relative to the reference field), for field sizes less than 1.0 cm. Conclusions: This work found that the Gafchromic{sup Registered-Sign} EBT2 film is adequate for small photon beam measurements, particularly for tissue maximum and off-axis ratios. However, careful attention must be taken when measuring output factors of small beams below 1.0 cm due to the film's energy dependence. The measurement differences may be attributable to the film's active layer composition because EBT2 incorporates higher Z elements (i.e., bromide and potassium), hence revealing a potential energy dependence for the dosimetry of small photon beams.

  3. Investigation of absorption spectra of Gafchromic EBT2 film's components and their impact on UVR dosimetry

    NASA Astrophysics Data System (ADS)

    Aydarous, Abdulkadir

    2016-05-01

    The absorption spectra of the EBT2 film's components were investigated in conjunction with its use for UVA dosimetry. The polyester (topside) and adhesive layers of the EBT2 film have been gently removed. Gafchromic™ EBT2 films with and without the protected layers (polyester and adhesive) were exposed to UVR of 365 nm for different durations. Thereafter, the UV-visible spectra were measured using a UV-visible spectrophotometer (Model Spectro Dual Split Beam, UVS-2700). Films were digitized using a Nikon CanoScan 9000F Mark II flatbed scanner. The dosimetric characteristics including film's uniformity, reproducibility and post-irradiation development were investigated. The color development of EBT2 and new modified EBT2 (EBT2-M) films irradiated with UVA was relatively stable (less than 1%) immediately after exposure. Based on this study, the sensitivity of EBT2 to UVR with wavelength between ~350 nm and ~390 nm can significantly be enhanced if the adhesive layer (~25 μm) is removed. The polyester layer plays almost no part on absorbing UVR with wavelength between ~320 nm and ~390 nm. Furthermore, various sensitivities for the EBT2-M film has been established depending on the wavelength of analysis.

  4. On the use of Gafchromic EBT3 films for validating a commercial electron Monte Carlo dose calculation algorithm

    NASA Astrophysics Data System (ADS)

    Chan, EuJin; Lydon, Jenny; Kron, Tomas

    2015-03-01

    This study aims to investigate the effects of oblique incidence, small field size and inhomogeneous media on the electron dose distribution, and to compare calculated (Elekta/CMS XiO) and measured results. All comparisons were done in terms of absolute dose. A new measuring method was developed for high resolution, absolute dose measurement of non-standard beams using Gafchromic® EBT3 film. A portable U-shaped holder was designed and constructed to hold EBT3 films vertically in a reproducible setup submerged in a water phantom. The experimental film method was verified with ionisation chamber measurements and agreed to within 2% or 1 mm. Agreement between XiO electron Monte Carlo (eMC) and EBT3 was within 2% or 2 mm for most standard fields and 3% or 3 mm for the non-standard fields. Larger differences were seen in the build-up region where XiO eMC overestimates dose by up to 10% for obliquely incident fields and underestimates the dose for small circular fields by up to 5% when compared to measurement. Calculations with inhomogeneous media mimicking ribs, lung and skull tissue placed at the side of the film in water agreed with measurement to within 3% or 3 mm. Gafchromic film in water proved to be a convenient high spatial resolution method to verify dose distributions from electrons in non-standard conditions including irradiation in inhomogeneous media.

  5. Characterizing the marker-dye correction for Gafchromic EBT2 film: A comparison of three analysis methods

    SciTech Connect

    McCaw, Travis J.; Micka, John A.; DeWerd, Larry A.

    2011-10-15

    Purpose: Gafchromic EBT2 film has a yellow marker dye incorporated into the active layer of the film that can be used to correct the film response for small variations in thickness. This work characterizes the effect of the marker-dye correction on the uniformity and uncertainty of dose measurements with EBT2 film. The effect of variations in time postexposure on the uniformity of EBT2 is also investigated. Methods: EBT2 films were used to measure the flatness of a {sup 60}Co field to provide a high-spatial resolution evaluation of the film uniformity. As a reference, the flatness of the {sup 60}Co field was also measured with Kodak EDR2 films. The EBT2 films were digitized with a flatbed document scanner 24, 48, and 72 h postexposure, and the images were analyzed using three methods: (1) the manufacturer-recommended marker-dye correction, (2) an in-house marker-dye correction, and (3) a net optical density (OD) measurement in the red color channel. The field flatness was calculated from orthogonal profiles through the center of the field using each analysis method, and the results were compared with the EDR2 measurements. Uncertainty was propagated through a dose calculation for each analysis method. The change in the measured field flatness for increasing times postexposure was also determined. Results: Both marker-dye correction methods improved the field flatness measured with EBT2 film relative to the net OD method, with a maximum improvement of 1% using the manufacturer-recommended correction. However, the manufacturer-recommended correction also resulted in a dose uncertainty an order of magnitude greater than the other two methods. The in-house marker-dye correction lowered the dose uncertainty relative to the net OD method. The measured field flatness did not exhibit any unidirectional change with increasing time postexposure and showed a maximum change of 0.3%. Conclusions: The marker dye in EBT2 can be used to improve the response uniformity of the film

  6. Accurate dosimetry with GafChromic EBT film of a 6 MV photon beam in water: What level is achievable?

    SciTech Connect

    Battum, L. J. van; Hoffmans, D.; Piersma, H.; Heukelom, S.

    2008-02-15

    This paper focuses on the accuracy, in absolute dose measurements, with GafChromic EBT film achievable in water for a 6 MV photon beam up to a dose of 2.3 Gy. Motivation is to get an absolute dose detection system to measure up dose distributions in a (water) phantom, to check dose calculations. An Epson 1680 color (red green blue) transmission flatbed scanner has been used as film scanning system, where the response in the red color channel has been extracted and used for the analyses. The influence of the flatbed film scanner on the film based dose detection process was investigated. The scan procedure has been optimized; i.e. for instance a lateral correction curve was derived to correct the scan value, up to 10%, as a function of optical density and lateral position. Sensitometric curves of different film batches were evaluated in portrait and landscape scan mode. Between various batches important variations in sensitometric curve were observed. Energy dependence of the film is negligible, while a slight variation in dose response is observed for very large angles between film surface and incident photon beam. Improved accuracy in absolute dose detection can be obtained by repetition of a film measurement to tackle at least the inherent presence of film inhomogeneous construction. We state that the overall uncertainty is random in absolute EBT film dose detection and of the order of 1.3% (1 SD) under the condition that the film is scanned in a limited centered area on the scanner and at least two films have been applied. At last we advise to check a new film batch on its characteristics compared to available information, before using that batch for absolute dose measurements.

  7. Variations in dose distribution and optical properties of Gafchromic{sup TM} EBT2 film according to scanning mode

    SciTech Connect

    Park, Soah; Kang, Sei-Kwon; Cheong, Kwang-Ho; Hwang, Taejin; Kim, Haeyoung; Han, Taejin; Lee, Me-Yeon; Kim, KyoungJu; Bae, Hoonsik; Su Kim, Hyeong; Han Kim, Jung; Jae Oh, Seung; Suh, Jin-Suck

    2012-05-15

    Purpose: The authors aim was to investigate the effects of using transmission and reflection scanning modes, the film orientation during scanning, and ambient room light on a dosimetry system based on the Gafchromic{sup TM} EBT2 film model. Methods: For calibration, the films were cut to 3 x 3 cm{sup 2} and irradiated from 20 to 700 cGy at the depth of maximum dose using 6 and 10 MV photon beams in a 10 x 10 cm{sup 2} field size. Absolute dose calibration of the linear accelerator was done according to the TRS398 protocol. An FG65-G ionization chamber was used to monitor the dose while irradiating the films in solid water. The film pieces were scanned with an EPSON Expression 1680 Pro flatbed scanner in transmission and reflection modes. Authors investigated the effect of orientation on films and examined the optical properties of EBT2 film using an ellipsometer and an ultraviolet (UV)/visible spectrometer to explain the dosimetric dependence of the film on orientation during the scanning process. To investigate the effect of ambient room light, films were preirradiated in 6 and 10 MV photon beams with intensity-modulated radiotherapy (IMRT) quality assurance (QA) plans, and then exposed to room light, either directly for 2 days in a workroom or for 2 months in a film box. Gamma index pass criteria of (3%, 3 mm) were used. Results: The dose response curves based on net optical density (NOD) indicated that the reflection scanning mode can provide a better dose sensitivity than the transmission scanning mode, whereas the standard deviation of the dose is greater in reflection mode than in transmission mode. When the film was rotated 90 deg. from the portrait orientation, the average dose of the EBT2 film decreased by 11.5-19.6% in transmission mode and by 1.5-2.3% in reflection mode. Using an ellipsometer, variation of the refractive index of EBT2 film--the birefringence property--was found to be the largest between 45 deg. (1.72 and 1.71) and 135 deg. (1.8 and 1

  8. Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies

    SciTech Connect

    Arjomandy, Bijan; Tailor, Ramesh; Anand, Aman; Sahoo, Narayan; Gillin, Michael; Prado, Karl; Vicic, Milos

    2010-05-15

    Purpose: Since the Gafchromic film EBT has been recently replaced by the newer model EBT2, its characterization, especially energy dependence, has become critically important. The energy dependence of the dose response of Gafchromic EBT2 film is evaluated for a broad range of energies from different radiation sources used in radiation therapy. Methods: The beams used for this study comprised of kilovoltage x rays (75, 125, and 250 kVp), {sup 137}Cs gamma (662 KeV), {sup 60}Co gamma (1.17-1.33 MeV), megavoltage x rays (6 and 18 MV), electron beams (6 and 20 MeV), and proton beams (100 and 250 MeV). The film's response to each of the above energies was measured over the dose range of 0.4-10 Gy, which corresponds to optical densities ranging from 0.05 to 0.74 for the film reader used. Results: The energy dependence of EBT2 was found to be relatively small within measurement uncertainties (1{sigma}={+-}4.5%) for all energies and modalities. Conclusion: For relative and absolute dosimetry of radiation therapy beams, the weak energy dependence of the EBT2 makes it most suitable for clinical use compared to other films.

  9. The use of new GAFCHROMIC EBT film for {sup 125}I seed dosimetry in Solid Water phantom

    SciTech Connect

    Chiu-Tsao, Sou-Tung; Medich, David; Munro, John III

    2008-08-15

    Radiochromic film dosimetry has been extensively used for intravascular brachytherapy applications for near field within 1 cm from the sources. With the recent introduction of new model of radiochromic films, GAFCHROMIC EBT, with higher sensitivity than earlier models, it is promising to extend the distances out to 5 cm for low dose rate (LDR) source dosimetry. In this study, the use of new model GAFCHROMIC EBT film for {sup 125}I seed dosimetry in Solid Water was evaluated for radial distances from 0.06 cm out to 5 cm. A multiple film technique was employed for four {sup 125}I seeds (Implant Sciences model 3500) with NIST traceable air kerma strengths. Each experimental film was positioned in contact with a {sup 125}I seed in a Solid Water phantom. The products of the air kerma strength and exposure time ranged from 8 to 3158 U-h, with the initial air kerma strength of 6 U in a series of 25 experiments. A set of 25 calibration films each was sequentially exposed to one {sup 125}I seed at about 0.58 cm distance for doses from 0.1 to 33 Gy. A CCD camera based microdensitometer, with interchangeable green (520 nm) and red (665 nm) light boxes, was used to scan all the films with 0.2 mm pixel resolution. The dose to each {sup 125}I calibration film center was calculated using the air kerma strength of the seed (incorporating decay), exposure time, distance from seed center to film center, and TG43U1S1 recommended dosimetric parameters. Based on the established calibration curve, dose conversion from net optical density was achieved for each light source. The dose rate constant was determined as 0.991 cGy U{sup -1} h{sup -1} ({+-}6.9%) and 1.014 cGy U{sup -1} h{sup -1} ({+-}6.8%) from films scanned using green and red light sources, respectively. The difference between these two values was within the uncertainty of the measurement. Radial dose function and 2D anisotropy function were also determined. The results obtained using the two light sources corroborated each

  10. In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment.

    PubMed

    Severgnini, Mara; de Denaro, Mario; Bortul, Marina; Vidali, Cristiana; Beorchia, Aulo

    2015-01-01

    Intraoperative electron radiation therapy (IOERT) cannot usually benefit, as conventional external radiotherapy, from software systems of treatment planning based on computed tomography and from common dose verify procedures. For this reason, in vivo film dosimetry (IVFD) proves to be an effective methodology to evaluate the actual radiation dose delivered to the target. A practical method for IVFD during breast IOERT was carried out to improve information on the dose actually delivered to the tumor target and on the alignment of the shielding disk with respect to the electron beam. Two EBT3 GAFCHROMIC films have been positioned on the two sides of the shielding disk in order to obtain the dose maps at the target and beyond the disk. Moreover the postprocessing analysis of the dose distribution measured on the films provides a quantitative estimate of the misalignment between the collimator and the disk. EBT3 radiochromic films have been demonstrated to be suitable dosimeters for IVD due to their linear dose-optical density response in a narrow range around the prescribed dose, as well as their capability to be fixed to the shielding disk without giving any distortion in the dose distribution. Off-line analysis of the radiochromic film allowed absolute dose measurements and this is indeed a very important verification of the correct exposure to the target organ, as well as an estimate of the dose to the healthy tissue underlying the shielding. These dose maps allow surgeons and radiation oncologists to take advantage of qualitative and quantitative feedback for setting more accurate treatment strategies and further optimized procedures. The proper alignment using elastic bands has improved the absolute dose accuracy and the collimator disk alignment by more than 50%. PMID:25679150

  11. GafChromic EBT film dosimetry with flatbed CCD scanner: A novel background correction method and full dose uncertainty analysis

    SciTech Connect

    Saur, Sigrun; Frengen, Jomar

    2008-07-15

    Film dosimetry using radiochromic EBT film in combination with a flatbed charge coupled device scanner is a useful method both for two-dimensional verification of intensity-modulated radiation treatment plans and for general quality assurance of treatment planning systems and linear accelerators. Unfortunately, the response over the scanner area is nonuniform, and when not corrected for, this results in a systematic error in the measured dose which is both dose and position dependent. In this study a novel method for background correction is presented. The method is based on the subtraction of a correction matrix, a matrix that is based on scans of films that are irradiated to nine dose levels in the range 0.08-2.93 Gy. Because the response of the film is dependent on the film's orientation with respect to the scanner, correction matrices for both landscape oriented and portrait oriented scans were made. In addition to the background correction method, a full dose uncertainty analysis of the film dosimetry procedure was performed. This analysis takes into account the fit uncertainty of the calibration curve, the variation in response for different film sheets, the nonuniformity after background correction, and the noise in the scanned films. The film analysis was performed for film pieces of size 16x16 cm, all with the same lot number, and all irradiations were done perpendicular onto the films. The results show that the 2-sigma dose uncertainty at 2 Gy is about 5% and 3.5% for landscape and portrait scans, respectively. The uncertainty gradually increases as the dose decreases, but at 1 Gy the 2-sigma dose uncertainty is still as good as 6% and 4% for landscape and portrait scans, respectively. The study shows that film dosimetry using GafChromic EBT film, an Epson Expression 1680 Professional scanner and a dedicated background correction technique gives precise and accurate results. For the purpose of dosimetric verification, the calculated dose distribution can

  12. A dosimetric study of small photon fields using polymer gel and Gafchromic EBT films

    SciTech Connect

    Hassani, Hossein; Nedaie, Hassan Ali; Zahmatkesh, Mohammad Hassan; Shirani, Kaveh

    2014-04-01

    The use of small field sizes is increasingly becoming important in radiotherapy particularly since the introduction of stereotactic radiosurgery and intensity-modulated radiation therapy techniques. The reliable measurement of delivered dose from such fields with conventional dosimeters, such as ionization chambers, is a challenging task. In this work, methacrylic and ascorbic acid in gelatin initiated by copper polymer gel dosimeters are employed to measure dose in 3 dimensions. Field sizes of 5 × 5 mm{sup 2}, 10 × 10 mm{sup 2}, 20 × 20 mm{sup 2}, and 30 × 30 mm{sup 2} are investigated for a 6-MV x-rays. The results show an agreement with Gafchromic film, with some variation in measured doses near the edge of the fields, where the film data decrease more rapidly than the other methods. Dose penumbra widths obtained with gel dosimeters and Gafchormic film were generally in agreement with each other. The results of this work indicate that polymer gel dosimetry could be invaluable for the quantification of the 3-dimensional dose distribution in small field size.

  13. Gafchromic EBT2 dosimetry via robust optimization

    NASA Astrophysics Data System (ADS)

    Alves, Victor G. L.; Cardoso, Simone C.; da Silva, Ademir X.

    2013-07-01

    An 'in house' software was developed with MATLAB in order to perform optimized dose calculations in radiotherapy. The aim of this work is to demonstrate an improvement on the Multichannel method using robust optimization that deals with optimization problems where robustness is sought against uncertainty. An optimization framework was developed in order to compare remaining error from optimization process of robust methods against the conventional triple-channel method. The proposed robust method minimizes the dose difference over all channels compared to the original triple-channel method, mainly over clinical dose range. Even if a Gafchromic EBT2 film irradiated by composite IMRT fields is analyzed, more consistent values than the ones obtained by the triple-channel method are found and Newton Rings patterns are minimized. When robust methods are applied, the difference between blue and red channel doses was found to be very small, about 104 times less than obtained by triple-channel optimization. It is well known that one outlier may cause a large error in a least squares estimator. The blue channel correction of non-uniformities may have best performance when robust optimization is used. A variety of anomalies (artifacts, Newton rings and other disturbances) behave differently from natural Gaussian random noise such as variations of the thickness. The use of robust optimization methods might be more realistic since this approach uses fatter tail distributions as the Laplace and could mitigate the Newton's Rings Pattern.

  14. Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film.

    PubMed

    Sonier, Marcus; Wronski, Matt; Yeboah, Collins

    2015-01-01

    Lens dose is a concern during the treatment of facial lesions with anterior electron beams. Lead shielding is routinely employed to reduce lens dose and minimize late complications. The purpose of this work is twofold: 1) to measure dose pro-files under large-area lead shielding at the lens depth for clinical electron energies via film dosimetry; and 2) to assess the accuracy of the Pinnacle treatment plan-ning system in calculating doses under lead shields. First, to simulate the clinical geometry, EBT3 film and 4 cm wide lead shields were incorporated into a Solid Water phantom. With the lead shield inside the phantom, the film was positioned at a depth of 0.7 cm below the lead, while a variable thickness of solid water, simulating bolus, was placed on top. This geometry was reproduced in Pinnacle to calculate dose profiles using the pencil beam electron algorithm. The measured and calculated dose profiles were normalized to the central-axis dose maximum in a homogeneous phantom with no lead shielding. The resulting measured profiles, functions of bolus thickness and incident electron energy, can be used to estimate the lens dose under various clinical scenarios. These profiles showed a minimum lead margin of 0.5 cm beyond the lens boundary is required to shield the lens to ≤ 10% of the dose maximum. Comparisons with Pinnacle showed a consistent overestimation of dose under the lead shield with discrepancies of ~ 25% occur-ring near the shield edge. This discrepancy was found to increase with electron energy and bolus thickness and decrease with distance from the lead edge. Thus, the Pinnacle electron algorithm is not recommended for estimating lens dose in this situation. The film measurements, however, allow for a reasonable estimate of lens dose from electron beams and for clinicians to assess the lead margin required to reduce the lens dose to an acceptable level. PMID:27074448

  15. Evaluation of GafChromic EBT prototype B for external beam dose verification

    SciTech Connect

    Todorovic, M.; Fischer, M.; Cremers, F.; Thom, E.; Schmidt, R.

    2006-05-15

    The capability of the new GafChromic EBT prototype B for external beam dose verification is investigated in this paper. First the general characteristics of this film (dose response, postirradiation coloration, influence of calibration field size) were derived using a flat-bed scanner. In the dose range from 0.1 to 8 Gy, the sensitivity of the EBT prototype B film is ten times higher than the response of the GafChromic HS, which so far was the GafChromic film with the highest sensitivity. Compared with the Kodak EDR2 film, the response of the EBT is higher by a factor of 3 in the dose range from 0.1 to 8 Gy. The GafChromic EBT almost does not show a temporal growth of the optical density and there is no influence of the chosen calibration field size on the dose response curve obtained from this data. A MatLab program was written to evaluate the two-dimensional dose distributions from treatment planning systems and GafChromic EBT film measurements. Verification of external beam therapy (SRT, IMRT) using the above-mentioned approach resulted in very small differences between the planned and the applied dose. The GafChromic EBT prototype B together with the flat-bed scanner and MatLab is a successful approach for making the advantages of the GafChromic films applicable for verification of external beam therapy.

  16. SU-E-T-637: Age and Batch Dependence of Gafchromic EBT Films in Photon and Proton Beam Dosimetry

    SciTech Connect

    Das, I; Akino, Y

    2014-06-15

    Purpose: Gafchrmoic films have undergone significant changes in characteristic over time reflected by HS, EBT, EBT2, EBT3 name. Interand intra- EBT film variability have been studied and found to be significant. However, age and lot/batch type have not been studied in various radiation beams that are investigated in this study. Methods: Thirteen sets of films; 2 EBT, 6 EBT2 and 5 EBT3 films with different lot number and expiration date were acquired. Films were cut longitudinally in 3 cm width and sandwiched between two solid water slabs that were placed in a water phantom to eliminate air gap. Each set of films were irradiated longitudinally at dmax with 6 and 15 MV photon beams as well as in reference condition (16 cm range, 10 cm SOBP) in our uniform scanning proton beam. Films were scanned using an Epson flatbed scanner (ES-10000G) after 48 hours to achieve full polymerization. The profiles were compared with the depth-dose measured with ionization chamber and net optical density (net OD) were calculated. Results: The net OD versus dose for EBT, EBT2 and EBT3 films of different age showed similar trend but with different slope. Even after calibration, differences are clearly visible in net OD in proton and photon beams. A net OD difference of nearly 0.5 is observed in photon but this was limited to 0.2–0.3 in proton beam. This relates to 20% and 15% dosimetric difference in photon and proton beam respectively over age and type of film. Conclusion: Net OD related to dose is dependent on the age and lot of the film in both photon and proton beams. It is concluded that before any set of film is used, a calibration film should be used for a meaningful dosimetry. The expired films showed larger OD variation compared to unexpired films.

  17. Quantitative analysis of dose distribution to determine optimal width of respiratory gating window using Gafchromic EBT2 film

    NASA Astrophysics Data System (ADS)

    Lee, Sung Hyun; Kim, Kum Bae; Kim, Mi-Sook; Yoo, Hyung-Jun; Park, Seungwoo; Jung, Haijo; Ji, Young Hoon; Yi, Chul-Young

    2013-02-01

    The purpose of this study was to determine the dependence of the dose distribution on the width of the respiratory gating window by using radiochromic Gafchromic EBT2 film. An in-house three-dimensional breathing simulator was used with a 4-s cycle and a 3-cm movement. The gamma index and the 50, 95, and 20-80% dose distributions were individually analyzed with regard to static, 100 (full motion), 60, 40, 30, 20, and 15% respiratory gating windows. In addition, dose differences based on the different extents of exposure were compared and analyzed along with total beam delivery time. Dose distributions became increasingly similar to the static value with decreasing respiratory gating window width. The extent differences from the static case for the low-dose region were not significant; neither were the extent differences for the high-dose region and 30, 20, and 15% gating windows (P = 0.388, 0.275, respectively). However, the 40% gating window showed a significant difference (P = 0.001). Moreover, the treatment time for the 30% gating window was reduced by more than half compared to that for the 15% gating window. Thus, the 30% window would be a reasonable choice for maximizing the range of the gating window while markedly decreasing the dose difference and the treatment time.

  18. Dose calculation of megavoltage IMRT using convolution kernels extracted from GafChromic EBT film-measured pencil beam profiles

    NASA Astrophysics Data System (ADS)

    Naik, Mehul S.

    Intensity-modulated radiation therapy (IMRT) is a 3D conformal radiation therapy technique that utilizes either a multileaf intensity-modulating collimator (MIMiC used with the NOMOS Peacock system) or a multileaf collimator (MLC) on a conventional linear accelerator for beam intensity modulation to afford increased conformity in dose distributions. Due to the high-dose gradient regions that are effectively created, particular emphasis should be placed in the accurate determination of pencil beam kernels that are utilized by pencil beam convolution algorithms employed by a number of commercial IMRT treatment planning systems (TPS). These kernels are determined from relatively large field dose profiles that are typically collected using an ion chamber during commissioning of the TPS, while recent studies have demonstrated improvements in dose calculation accuracy when incorporating film data into the commissioning measurements. For this study, it has been proposed that the shape of high-resolution dose kernels can be extracted directly from single pencil beam (beamlet) profile measurements acquired using high-precision dosimetric film in order to accurately compute dose distributions, specifically for small fields and the penumbra regions of the larger fields. The effectiveness of GafChromic EBT film as an appropriate dosimeter to acquire the necessary measurements was evaluated and compared to the conventional silver-halide Kodak EDR2 film. Using the NOMOS Peacock system, similar dose kernels were extracted through deconvolution of the elementary pencil beam profiles using the two different types of films. Independent convolution-based calculations were performed using these kernels, resulting in better agreement with the measured relative dose profiles, as compared to those determined by CORVUS TPS' finite-size pencil beam (FSPB) algorithm. Preliminary evaluation of the proposed method in performing kernel extraction for an MLC-based IMRT system also showed

  19. SU-E-T-30: Absorbed Doses Determined by Texture Analysis of Gafchromic EBT3 Films Using Scanning Electron Microscopy: A Feasibility Study

    SciTech Connect

    Park, S; Kim, H; Ye, S

    2014-06-01

    Purpose: The texture analysis method is useful to estimate structural features of images as color, size, and shape. The study aims to determine a dose-response curve by texture analysis of Gafchromic EBT3 film images using scanning electron microscopy (SEM). Methods: The uncoated Gafchromic EBT3 films were prepared to directly scan over the active surface layer of EBT3 film using SEM. The EBT3 films were exposed at a dose range of 0 to 10 Gy using a 6 MV photon beam. The exposed film samples were SEM-scanned at 100X, 1000X, and 3000X magnifications. The four texture features (Homogeneity, Correlation, Contrast, and Energy) were calculated based on the gray level co-occurrence matrix (GLCM) derived from the SEM images at each dose. To validate a correlation between delivered doses and texture features, an R-squared value in linear regression was tested. Results: The results showed that the Correlation index was more suitable as dose indices than the other three texture features due to higher linearity and sensitivity of the dose response curves. Further the Correlation index of 3000X magnified SEM images with 9 pixel offsets had an R-squared value of 0.964. The differences between the delivered doses and the doses measured by this method were 0.9, 1.2, 0.2, and 0.2 Gy at 5, 10, 15, and 20 Gy, respectively. Conclusion: It seems to be feasible to convert micro-scale structural features of {sub χ}t{sub χχχ}he EBT3 films to absorbed doses using the texture analysis method.

  20. An investigation of backscatter factors for kilovoltage x-rays: a comparison between Monte Carlo simulations and Gafchromic EBT film measurements

    NASA Astrophysics Data System (ADS)

    Kim, J.; Hill, R.; Claridge Mackonis, E.; Kuncic, Z.

    2010-02-01

    Backscatter factors are important parameters in the determination of dose for kilovoltage x-ray beams. However, backscatter factors are difficult to measure experimentally, and tabulated values are based largely on Monte Carlo calculations. In this study we have determined new backscatter factors by both experimental and Monte Carlo methods, and compared them with existing backscatter factors published in the AAPM TG-61 protocol. The purpose of this study is twofold: (1) to evaluate the overall effectiveness of using Gafchromic EBT film for backscatter factor measurements and (2) to determine whether existing Monte Carlo-calculated backscatter factors need to be updated. We measured backscatter factors using Gafchromic EBT film for three field sizes (2, 4 and 6 cm diameter cones) and three kilovoltage beam qualities, including 280 kVp for which similar measurements have not previously been reported. We also present new Monte Carlo-calculated backscatter factors obtained using the EGSnrc/BEAMnrc code system to simulate the Pantak kilovoltage x-ray unit used in our measurements. The results were compared with backscatter factors tabulated in the AAPM TG-61 protocol for kilovoltage x-ray dosimetry. The largest difference between our measured and calculated backscatter factors and the AAPM TG-61 values was found to be 2.5%. This agreement is remarkably good, considering that the AAPM TG-61 values consist of a combination of experimental and Monte Carlo calculations obtained over 20 years ago using different measurement techniques, as well as older Monte Carlo code and cross-section data. Furthermore, our Monte Carlo-calculated backscatter factors agree within 1% with the AAPM TG-61 values for all beam qualities and field sizes. Our Gafchromic film measurements had slightly larger differences with the AAPM TG-61 backscatter factors, up to approximately 2% for the 6 cm diameter cone at a beam quality of 50 kVp. The largest difference in backscatter factors, of 2.5%, was

  1. Comparison of dosimeter response: ionization chamber, TLD, and Gafchromic EBT2 film in 3D-CRT, IMRT, and SBRT techniques for lung cancer

    NASA Astrophysics Data System (ADS)

    Fitriandini, A.; Wibowo, W. E.; Pawiro, S. A.

    2016-03-01

    This research was conducted by measuring point dose in the target area (lungs), heart, and spine using four dosimeters (PTW N30013, Exradin A16, TLD, and the Gafchromic EBT2 film). The measurement was performed in CIRS 002LFC thorax phantom. The main objective of this study was to compare the dosimetry of those different systems. Dose measurements performed only in a single fraction of irradiation. The measurements result shown that TLD has the least accuracy and precision. As the effect of volume averaging, ionization chamber reaches the discrepancy value up to -13.30% in the target area. EBT2 film has discrepancy value of <1% in the 3D-CRT and IMRT techniques. This dosimeter is proposed to be an appropriate alternative dosimeter to be used at point dose verification.

  2. TH-C-19A-04: Commissioning and Validation of EBT3 Gafchromic Films for Measurements of Dose Profiles and Integrals of Dose in Small Fields

    SciTech Connect

    Dufreneix, S; Bordy, J; Delaunay, F; Daures, J; Gouriou, J; Le, Roy M; Ostrowsky, A; Rapp, B; Sommier, L

    2014-06-15

    Purpose: The use of a primary dosimeter larger than the radiation field gives access to the integral of dose over a specified surface normal to the beam. If a relative dose profile of the beam is well known, it is then possible to calculate the distribution of the absorbed dose at any point on the considered surface. This study aims at validating the use of EBT3 gafchromic films for the measurement of 2D dose distribution and integrals of dose in small fields for such use. Methods: New EBT3 films have been fully characterized: the response versus energy, dose-rate and dose has been investigated. Profiles measured in circular field with a diameter of 20 mm have been compared to the ones measured with a diamond detector developed at CEA/LIST/LCD. The ratio of dose area products measured with EBT3 on a 6 mm and 30 mm diameter surface has been compared to the ratio measured with primary dosimeters (calorimeters) and calculated with Monte Carlo simulations. Results: There was no significant difference between the dose-calibration curves in a 6 MV and a 60Co beam. Deviation was within uncertainty bars when the dose rate inside a pulse was divided by a factor of 80 in the 6 MV photon beam. Profiles in small fields are in good agreement with the diamond profiles. Dose area product ratios obtained with EBT3, calorimeters and simulations are within 1%. Conclusion: EBT3 films are good candidates for the measurement of relative dose distribution in small fields as long as the average of several films is considered. They can be used in association with primary measurements to determinate dosimetric references in small fields and to transfer them to the end user.

  3. Feasibility of EBT Gafchromic films for comparison exercises among standard beta radiation fields.

    PubMed

    Benavente, J A; Meira-Belo, L C; Reynaldo, S R; da Silva, T A

    2012-12-01

    The feasibility of using radiochromic films to verify the metrological coherence among standard beta radiation fields was evaluated. Exercises were done between two Brazilian metrology laboratories in beta fields from (90)Sr/(90)Y, (85)Kr and (147)Pm radiation sources. Results showed that the radiochromic film was useful for field mapping aiming uniformity and alignment verification and it was not reliable for absorbed dose measurements only for (147)Pm beta field. PMID:22917942

  4. SU-E-T-121: Dosimetric Characterization of Gafchromic Film EBT3 Using Vidar DosimetryPro Advantage RED and EPSON Expression 10000XL Scanners

    SciTech Connect

    Medina, L; Adrada, A; Filipuzzi, M; Garrigo, E; Venencia, C

    2014-06-01

    Purpose: The purpose of this paper is to characterize EBT3 using two types of scanner, analyzing the factors of influence of each dosimetry system. Methods: The film used in this study was GAFCHROMIC EBT3, the films were exposed at a dose range between 0Gy a 9Gy in a solid water phantom, SSD=100cm, 5cm depth and perpendicularly to the 6MV photon beam generated by a Novalis TX linear accelerator equipped with an HDMLC. A Farmer type ion chamber TN30013 (PTW) was used to determine the dose delivered to the film. The films were digitized with a scanner EPSON expression 10000XL and the VIDAR DosimetryPro Adventage RED. Software RIT113v6.1 was used for construction of the calibration curve and analysis. The film characteristics investigated were: response at different dose levels, sensitivity to orientation and side and resolution through the results of the spatial response function by analyzing a step pattern. Additionally, 20 IMRT treatment fields were measured with both scanner and compared with calculated dose using gamma index analysis (3%-3mm). Results: The OD obtained for dose level 2Gy in the orientation portrait of the film on the scanner EPSON is (0,222±0,19) and for Vidar RED (0,252±0,10) and landscape is for EPSON (0,211±0,25) and for Vidar RED (0,250±0,11) . The orientation dependence with respect to film side is about 0,09% for EPSON and about 0.03% for VIDAR. The spatial response function increase in response to the Gaussian function FWHM EPSON scanner (0.18mm) compared with VIDAR scanner function (less than 0.06mm) was observed. We analyzed 20 total plan dose distributions the number of pixels with gamma>1 (3%-3mm) was 0.7%±1.2 [0.1%; 2.82%] for EBT3-VIDAR y 2%±2.9 [0.2%; 3.5%] for EBT3-EPSON. Conclusion: VIDAR scanner shows better sensitivity. EBT3 film shows a different response between portrait and landscape orientation. Step pattern is better reproduce by VIDAR scanner.

  5. SU-E-T-269: Quality Assurance of Spine Volumetric Modulated Arc Therapy with Flattening Filter Free Beams Using Gafchromic EBT3 Films

    SciTech Connect

    Choi, Y; Cho, B; Kwak, J; Ahn, S; Choi, E; Kim, J; Song, S; Yoon, S; Kim, Y; Kim, S; Park, J

    2014-06-01

    Purpose: We implemented the Gafchromic film-based patient specific QA of volumetric modulated arc therapy (VMAT) with flattening-filter free (FFF) beams for spine metastases and validated the accuracy of fast arc delivery. Methods: EBT3 films and a homemade cylindrical QA phantom were employed for dosimetric verification of VMATs. For 14 FFF VMAT plans (10 with 10-MV FFF beams and 4 with 6-MV FFF beams), the doses were recalculated on the phantom and delivered by a TrueBeam STx accelerator equipped with a high-definition 120 leaf MLC. The EBT3 films were scanned using an Epson 10000XL scanner through the FilmQA Pro software. All the irradiated film images were converted to dose map using a calibration response curve. The resulting dose map of film measurement was compared with treatment plan and evaluated using gamma analysis with dose tolerance of 2% within 2 mm. In addition, the point-dose measurement in the phantom using an ion chamber was evaluated as a reference in a ratio of measured and planned doses. Results: The gamma pass rates averaged over all FFF plans for composite-field measurements were 96.0 ± 3.6% (88.9%–99.5%). When adopting a tolerance level of 3% - 3 mm, the gamma pass rates were improved with the ranges from 98% to 100%. In addition, dose profiles and dose distributions showed that spinal cord was protected by the rapid dose fall-off and by delivering the treatment with high precision. In point-dose measurements, the average differences between the measured and planned doses were 0.5% ± 1.0% of the prescription dose. Conclusion: We demonstrated that Gafchromic EBT3 film would be an effective patient-specific QA tool, especially for VMAT of spine SBRT with treatment of small fields and highly gradient dose distributions. The results of film QA verified that the dosimetric accuracy of spine SBRT utilizing RapidArc with FFF beams in our institution is reliable.

  6. Comparison of Gafchromic EBT2 and EBT3 for patient-specific quality assurance: Cranial stereotactic radiosurgery using volumetric modulated arc therapy with multiple noncoplanar arcs

    SciTech Connect

    Fiandra, Christian; Fusella, Marco; Filippi, Andrea Riccardo; Ricardi, Umberto; Ragona, Riccardo; Giglioli, Francesca Romana; Mantovani, Cristina

    2013-08-15

    Purpose: Patient-specific quality assurance in volumetric modulated arc therapy (VMAT) brain stereotactic radiosurgery raises specific issues on dosimetric procedures, mainly represented by the small radiation fields associated with the lack of lateral electronic equilibrium, the need of small detectors and the high dose delivered (up to 30 Gy). Gafchromic{sup TM} EBT2 and EBT3 films may be considered the dosimeter of choice, and the authors here provide some additional data about uniformity correction for this new generation of radiochromic films.Methods: A new analysis method using blue channel for marker dye correction was proposed for uniformity correction both for EBT2 and EBT3 films. Symmetry, flatness, and field-width of a reference field were analyzed to provide an evaluation in a high-spatial resolution of the film uniformity for EBT3. Absolute doses were compared with thermoluminescent dosimeters (TLD) as baseline. VMAT plans with multiple noncoplanar arcs were generated with a treatment planning system on a selected pool of eleven patients with cranial lesions and then recalculated on a water-equivalent plastic phantom by Monte Carlo algorithm for patient-specific QA. 2D quantitative dose comparison parameters were calculated, for the computed and measured dose distributions, and tested for statistically significant differences.Results: Sensitometric curves showed a different behavior above dose of 5 Gy for EBT2 and EBT3 films; with the use of inhouse marker-dye correction method, the authors obtained values of 2.5% for flatness, 1.5% of symmetry, and a field width of 4.8 cm for a 5 × 5 cm{sup 2} reference field. Compared with TLD and selecting a 5% dose tolerance, the percentage of points with ICRU index below 1 was 100% for EBT2 and 83% for EBT3. Patients analysis revealed statistically significant differences (p < 0.05) between EBT2 and EBT3 in the percentage of points with gamma values <1 (p= 0.009 and p= 0.016); the percent difference as well as

  7. Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films.

    PubMed

    Dreindl, Ralf; Georg, Dietmar; Stock, Markus

    2014-05-01

    Gafchromic® EBT2 film is a widely used dosimetric tool for quality assurance in radiation therapy. In 2012 EBT3 was presented as a replacement for EBT2 films. The symmetric structure of EBT3 films to reduce face-up/down dependency as well as the inclusion of a matte film surface to frustrate Newton Ring artifacts present the most prominent improvements of EBT3 films. The aim of this study was to investigate the characteristics of EBT3 films, to benchmark the films against the known EBT2-features and to evaluate the dosimetric behavior over a time period greater than 6 months. All films were irradiated to clinical photon beams (6 MV, 10 MV and 18 MV) on an Elekta Synergy Linac equipped with a Beam Modulator MLC in solid water phantom slabs. Film digitalization was done with a flatbed transparency scanner (Type Epson Expression 1680 Pro). MATLAB® was used for further statistical calculations and image processing. The investigations on post-irradiation darkening, film orientation, film uniformity and energy dependency resulted in negligible differences between EBT2 and EBT3 film. A minimal improvement in face-up/down dependence was found for EBT3. The matte film surface of EBT3 films turned out to be a practical feature as Newton rings could be eliminated completely. Considering long-term behavior (> 6 months) a shift of the calibration curve for EBT2 and EBT3 films due to changes in the dynamic response of the active component was observed. In conclusion, the new EBT3 film yields comparable results to its predecessor EBT2. The general advantages of radiochromic film dosimeters are completed by high film homogeneity, low energy dependence for the observed energy range and a minimized face-up/down dependence. EBT2 dosimetry-protocols can also be used for EBT3 films, but the inclusion of periodical recalibration-interval (e.g. once a quarter) is recommended for protocols of both film generations. PMID:24055395

  8. Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge-doped optical fiber, EBT3 Gafchromic film, and PRESAGE{sup Registered-Sign} radiochromic plastic

    SciTech Connect

    Palmer, A. L.; Di Pietro, P.; Alobaidli, S.; Issa, F.; Doran, S.; Bradley, D.; Nisbet, A.

    2013-06-15

    Purpose: Dose distribution measurement in clinical high dose rate (HDR) brachytherapy is challenging, because of the high dose gradients, large dose variations, and small scale, but it is essential to verify accurate treatment planning and treatment equipment performance. The authors compare and evaluate three dosimetry systems for potential use in brachytherapy dose distribution measurement: Ge-doped optical fibers, EBT3 Gafchromic film with multichannel analysis, and the radiochromic material PRESAGE{sup Registered-Sign} with optical-CT readout. Methods: Ge-doped SiO{sub 2} fibers with 6 {mu}m active core and 5.0 mm length were sensitivity-batched and their thermoluminescent properties used via conventional heating and annealing cycles. EBT3 Gafchromic film of 30 {mu}m active thickness was calibrated in three color channels using a nominal 6 MV linear accelerator. A 48-bit transmission scanner and advanced multichannel analysis method were utilized to derive dose measurements. Samples of the solid radiochromic polymer PRESAGE{sup Registered-Sign }, 60 mm diameter and 100 mm height, were analyzed with a parallel beam optical CT scanner. Each dosimetry system was used to measure the dose as a function of radial distance from a Co-60 HDR source, with results compared to Monte Carlo TG-43 model data. Each system was then used to measure the dose distribution along one or more lines through typical clinical dose distributions for cervix brachytherapy, with results compared to treatment planning system (TPS) calculations. Purpose-designed test objects constructed of Solid Water and held within a full-scatter water tank were utilized. Results: All three dosimetry systems reproduced the general shape of the isolated source radial dose function and the TPS dose distribution. However, the dynamic range of EBT3 exceeded those of doped optical fibers and PRESAGE{sup Registered-Sign }, and the latter two suffered from unacceptable noise and artifact. For the experimental

  9. Dosimetric verification of stereotactic radiosurgery/stereotactic radiotherapy dose distributions using Gafchromic EBT3

    SciTech Connect

    Cusumano, Davide; Fumagalli, Maria L.; Marchetti, Marcello; Fariselli, Laura; De Martin, Elena

    2015-10-01

    Aim of this study is to examine the feasibility of using the new Gafchromic EBT3 film in a high-dose stereotactic radiosurgery and radiotherapy quality assurance procedure. Owing to the reduced dimensions of the involved lesions, the feasibility of scanning plan verification films on the scanner plate area with the best uniformity rather than using a correction mask was evaluated. For this purpose, signal values dispersion and reproducibility of film scans were investigated. Uniformity was then quantified in the selected area and was found to be within 1.5% for doses up to 8 Gy. A high-dose threshold level for analyses using this procedure was established evaluating the sensitivity of the irradiated films. Sensitivity was found to be of the order of centiGray for doses up to 6.2 Gy and decreasing for higher doses. The obtained results were used to implement a procedure comparing dose distributions delivered with a CyberKnife system to planned ones. The procedure was validated through single beam irradiation on a Gafchromic film. The agreement between dose distributions was then evaluated for 13 patients (brain lesions, 5 Gy/die prescription isodose ~80%) using gamma analysis. Results obtained using Gamma test criteria of 5%/1 mm show a pass rate of 94.3%. Gamma frequency parameters calculation for EBT3 films showed to strongly depend on subtraction of unexposed film pixel values from irradiated ones. In the framework of the described dosimetric procedure, EBT3 films proved to be effective in the verification of high doses delivered to lesions with complex shapes and adjacent to organs at risk.

  10. Absolute dosimetric characterization of Gafchromic EBT3 and HDv2 films using commercial flat-bed scanners and evaluation of the scanner response function variability

    NASA Astrophysics Data System (ADS)

    Chen, S. N.; Gauthier, M.; Bazalova-Carter, M.; Bolanos, S.; Glenzer, S.; Riquier, R.; Revet, G.; Antici, P.; Morabito, A.; Propp, A.; Starodubtsev, M.; Fuchs, J.

    2016-07-01

    Radiochromic films (RCF) are commonly used in dosimetry for a wide range of radiation sources (electrons, protons, and photons) for medical, industrial, and scientific applications. They are multi-layered, which includes plastic substrate layers and sensitive layers that incorporate a radiation-sensitive dye. Quantitative dose can be retrieved by digitizing the film, provided that a prior calibration exists. Here, to calibrate the newly developed EBT3 and HDv2 RCFs from Gafchromic™, we used the Stanford Medical LINAC to deposit in the films various doses of 10 MeV photons, and by scanning the films using three independent EPSON Precision 2450 scanners, three independent EPSON V750 scanners, and two independent EPSON 11000XL scanners. The films were scanned in separate RGB channels, as well as in black and white, and film orientation was varied. We found that the green channel of the RGB scan and the grayscale channel are in fact quite consistent over the different models of the scanner, although this comes at the cost of a reduction in sensitivity (by a factor ˜2.5 compared to the red channel). To allow any user to extend the absolute calibration reported here to any other scanner, we furthermore provide a calibration curve of the EPSON 2450 scanner based on absolutely calibrated, commercially available, optical density filters.

  11. Absolute dosimetric characterization of Gafchromic EBT3 and HDv2 films using commercial flat-bed scanners and evaluation of the scanner response function variability.

    PubMed

    Chen, S N; Gauthier, M; Bazalova-Carter, M; Bolanos, S; Glenzer, S; Riquier, R; Revet, G; Antici, P; Morabito, A; Propp, A; Starodubtsev, M; Fuchs, J

    2016-07-01

    Radiochromic films (RCF) are commonly used in dosimetry for a wide range of radiation sources (electrons, protons, and photons) for medical, industrial, and scientific applications. They are multi-layered, which includes plastic substrate layers and sensitive layers that incorporate a radiation-sensitive dye. Quantitative dose can be retrieved by digitizing the film, provided that a prior calibration exists. Here, to calibrate the newly developed EBT3 and HDv2 RCFs from Gafchromic™, we used the Stanford Medical LINAC to deposit in the films various doses of 10 MeV photons, and by scanning the films using three independent EPSON Precision 2450 scanners, three independent EPSON V750 scanners, and two independent EPSON 11000XL scanners. The films were scanned in separate RGB channels, as well as in black and white, and film orientation was varied. We found that the green channel of the RGB scan and the grayscale channel are in fact quite consistent over the different models of the scanner, although this comes at the cost of a reduction in sensitivity (by a factor ∼2.5 compared to the red channel). To allow any user to extend the absolute calibration reported here to any other scanner, we furthermore provide a calibration curve of the EPSON 2450 scanner based on absolutely calibrated, commercially available, optical density filters. PMID:27475550

  12. Poster — Thur Eve — 26: Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film

    SciTech Connect

    Sanier, M; Wronski, M; Yeboah, C

    2014-08-15

    The purpose of this work is twofold: 1) to measure dose profiles under lead shielding at the level of the lens for a range of clinical electron energies via film dosimetry; and, 2) to assess the validity of the Pinnacle treatment planning system (TPS) in calculating the penumbral doses under lead shielding with the heterogeneous electron algorithm. First, a film calibration curve that spanned the electron energies of interest, 6–18MeV, was created. Next, EBT3 film and lead shielding were incorporated into a solid water phantom with the film positioned 7mm below the lead and a variable thickness of bolus on top. This geometry was reproduced in the Pinnacle TPS and used to calculate dose profiles using the heterogeneous electron algorithm. The measured vs. calculated dose profiles were normalized to d{sub max} in a homogeneous phantom with no lead shielding and compared. Pinnacle consistently overestimated the dose distal to the lead shielding with significant discrepancies occurring near the edge of the lead shield reaching 25% at the edge and 35% in the open field region. The film measurements showed that a minimum lead margin of 5mm extending beyond the diameter of the lens is required to adequately shield the lens to ≤10% of the dose at d{sub max}. These measurements allow for a reasonable estimate of the dose to the lens from anterior electron beams. They also allow for clinicians to assess the extent of the lead margin required to reduce the lens dose to an acceptable amount prior to radiotherapy treatment.

  13. Characterization of the effect of MRI on Gafchromic film dosimetry.

    PubMed

    Reyhan, Meral L; Chen, Ting; Zhang, Miao

    2015-01-01

    Magnetic resonance (MR) imaging of Gafchromic film causes perturbation to absolute dosimetry measurements; the purpose of this work was to characterize the perturbation and develop a correction method for it. Three sets of Gafchromic EBT2 film were compared: radiation (control), radiation followed by MR imaging (RAD + B), and MR imaging followed by radiation (B + RAD). The T1-weighted and T2-weighted MR imaging was performed using a 1.5T scanner with the films wedged between two chicken legs. Doses from 0 to 800 cGy were delivered with a 6MV linac. The time interval between radiation and MR imaging was less than 10 min. Film calibration was generated from the red channel. Microscopic imaging was performed on two pieces of film. The effect of specific absorption rate (SAR) was determined by exposing another three sets of films to low, medium, and high levels of SAR through a series of pulse sequences. No discernible preferential alignment was detected on the microscopic images of the irradiated film exposed to MRI. No imaging artifacts were introduced by Gafchromic film on any MR images. On average, 4% dose difference was observed between B + RAD or RAD + B and the control, using the same calibration curve. The pixel values between the B + RAD or RAD + B and the control films were found to follow a linear relationship pixel(Control) = 1.02 × pixel(B + RAD or RAD + B). By applying this correction, the average dose error was reduced to approximately 2%. The SAR experiment revealed a dose overestimation with increasing SAR even when the correction was applied. It was concluded that MR imaging introduces perturbation on Gafchromic film dose measurements by 4% on average, compared to calibrating the film without the presence of MRI. This perturbation can be corrected by applying a linear correction to the pixel values. Additionally, Gafchromic film did not introduce any imaging artifacts in any of the MR images acquired. PMID:26699587

  14. Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams

    SciTech Connect

    Brown, Thomas A. D.; Hogstrom, Kenneth R.; Alvarez, Diane; Matthews, Kenneth L. II; Ham, Kyungmin; Dugas, Joseph P.

    2012-12-15

    Purpose: This work investigates the dose-response curves of GAFCHROMIC{sup Registered-Sign} EBT, EBT2, and EBT3 radiochromic films using synchrotron-produced monochromatic x-ray beams. EBT2 film is being utilized for dose verification in photoactivated Auger electron therapy at the Louisiana State University Center for Advanced Microstructures and Devices (CAMD) synchrotron facility. Methods: Monochromatic beams of 25, 30, and 35 keV were generated on the tomography beamline at CAMD. Ion chamber depth-dose measurements were used to determine the dose delivered to films irradiated at depths from 0.7 to 8.5 cm in a 10 Multiplication-Sign 10 Multiplication-Sign 10-cm{sup 3} polymethylmethacrylate phantom. AAPM TG-61 protocol was applied to convert measured ionization into dose. Films were digitized using an Epson 1680 Professional flatbed scanner and analyzed using the net optical density (NOD) derived from the red channel. A dose-response curve was obtained at 35 keV for EBT film, and at 25, 30, and 35 keV for EBT2 and EBT3 films. Calibrations of films for 4 MV x-rays were obtained for comparison using a radiotherapy accelerator at Mary Bird Perkins Cancer Center. Results: The sensitivity (NOD per unit dose) of EBT film at 35 keV relative to that for 4-MV x-rays was 0.73 and 0.76 for doses 50 and 100 cGy, respectively. The sensitivity of EBT2 film at 25, 30, and 35 keV relative to that for 4-MV x-rays varied from 1.09-1.07, 1.23-1.17, and 1.27-1.19 for doses 50-200 cGy, respectively. For EBT3 film the relative sensitivity was within 3% of unity for all three monochromatic x-ray beams. Conclusions: EBT and EBT2 film sensitivity showed strong energy dependence over an energy range of 25 keV-4 MV, although this dependence becomes weaker for larger doses. EBT3 film shows weak energy dependence, indicating that it would be a better dosimeter for kV x-ray beams where beam hardening effects can result in large changes in the effective energy.

  15. Measuring solar UV radiation with EBT radiochromic film.

    PubMed

    Butson, Ethan T; Cheung, Tsang; Yu, Peter K N; Butson, Martin J

    2010-10-21

    Ultraviolet radiation dosimetry has been performed with the use of a radiochromic film dosimeter called Gafchromic EBT for solar radiation exposure. The film changes from a clear colour to blue colour when exposed to ultraviolet radiation and results have shown that the colour change is reproducible within ±10% at 5 kJ m(-2) UV exposure under various conditions of solar radiation. Parameters tested included changes in season (summer versus winter exposure), time of day, as well as sky conditions such as cloudy skies versus clear skies. As the radiochromic films' permanent colour change occurs in the visible wavelengths the film can be analysed with a desktop scanner with the most sensitive channel for analysis being the red component of the signal. Results showed that an exposure of 5 kJ m(-2) (approximately 1 h exposure in full sun during summer) produced an approximate 0.28 change in the net OD when analysed in reflection mode on the desktop scanner which is significant darkening. The main advantages of this film type, and thus the new EBT2 film which has replaced EBT for measurement of UV exposure, is the visible colour change and thus easy analysis using a desktop scanner, its uniformity in response and its robust physical strength for use in outside exposure situations. PMID:20858922

  16. Comparison of EBT and EBT3 RadioChromic Film Usage in Parotid Cancer Radiotherapy

    PubMed Central

    Bahreyni Toossi, M.T.; Khorshidi, F.; Ghorbani, M.; Mohamadian, N.; Davenport, D.

    2016-01-01

    Background EBT and EBT3 radioChromic films have been used in radiotherapy dosimetry for years. Objective The aim of the current study is to compare EBT and EBT3 radioChromic films in dosimetry of radiotherapy fields for treatment of parotid cancer. Methods Calibrations of EBT and EBT3 films were performed with identical setups using a 6 MV photon beam of a Siemens Primus linac. Skin dose was measured at different points in the right anterior oblique (RAO) and right posterior oblique (RPO) fields by EBT and EBT3 films on a RANDO phantom. Results While dosimetry was performed with the same conditions for the two film types for calibration and in phantom in parotid cancer radiotherapy, the measured net optical density (NOD) in EBT film was found to be higher than that from EBT3 film. The minimum difference between these two films under calibration conditions was about 2.9% (for 0.2 Gy) with a maximum difference of 35.5% (for 0.5 Gy). In the therapeutic fields of parotid cancer radiotherapy at different points, the measured dose from EBT film was higher than the EBT3 film. In these fields the minimum and maximum measured dose differences were 16.0% and 25.5%, respectively. Conclusion EBT film demonstrates higher NOD than EBT3 film. This effect may be related to the higher sensitivity of EBT film over EBT3 film. However, the obtained dose differences between these two films in low dose range can be due to the differences in fitting functions applied following the calibration process. PMID:27026949

  17. Measurement of low-energy backscatter factors using GAFCHROMIC film and OSLDs.

    PubMed

    Mart, Chris J; Elson, Howard R; Lamba, Michael A S

    2012-01-01

    Some of the lowest voltages used in radiotherapy are termed Grenz and superficial X-rays of ~ 20 and ~ 100 kVp, respectively. Dosimetrically, the surface doses from these beams are calculated with the use of a free in-air air kerma measurement combined with a backscatter factor and the appropriate ratio of mass energy absorption coefficients from the measurement material to water. Alternative tools to the standard ion chamber for measuring the BSF are GAFCHROMIC EBT2 film and optically stimulated luminescent dosimeter (OSLD) crystals made from Al2O3. The scope of this project included making three different backscatter measurements with an Xstrahl-D3100 X-ray unit on the Grenz ray and superficial settings. These measurements were with OSLDs, GAFCHROMIC EBT2 film, and a PTW ionization chamber. The varied measurement methods allowed for intercomparison to determine the accuracy of the results. The ion chamber measurement was the least accurate, as expected from previous experimental findings. GAFCHROMIC EBT2 film proved to be a useful tool which gave reasonable results, and Landauer OSLDs showed good results for smaller field sizes and an increasing overresponse with larger fields. The specific backscatter factors for this machine demonstrated values about 5% higher than the universal values suggested by the AAPM and IPEMB codes of practice for the 100 kVp setting. The 20 kvp measured data from both techniques showed general agreement with those found in the BJR Supplement No. 10, indicating that this unit's Grenz ray spectrum is similar to those used in previous experimental work. PMID:23149776

  18. NOTE: Absorption spectra variations of EBT radiochromic film from radiation exposure

    NASA Astrophysics Data System (ADS)

    Butson, M. J.; Cheung, T.; Yu, P. K. N.

    2005-07-01

    Gafchromic EBT radiochromic film is one of the newest radiation-induced auto-developing x-ray analysis films available for therapeutic radiation dosimetry in radiotherapy applications. The spectral absorption properties in the visible wavelengths have been investigated and results show two main peaks in absorption located at 636 nm and 585 nm. These absorption peaks are different to many other radiochromic film products such as Gafchromic MD-55 and HS film where two peaks were located at 676 nm and 617 nm respectively. The general shape of the absorption spectra is similar to older designs. A much higher sensitivity is found at high-energy x-rays with an average 0.6 OD per Gy variation in OD seen within the first Gy measured at 636 nm using 6 MV x-rays. This is compared to approximately 0.09 OD units for the first Gy at the 676 nm absorption peak for HS film at 6 MV x-ray energy. The film's blue colour is visually different from older varieties of Gafchromic film with a higher intensity of mid-range blue within the film. The film provides adequate relative absorbed dose measurement for clinical radiotherapy x-ray assessment in the 1 2 Gy dose range which with further investigation may be useful for fractionated radiotherapy dose assessment.

  19. Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapy

    NASA Astrophysics Data System (ADS)

    Richley, L.; John, A. C.; Coomber, H.; Fletcher, S.

    2010-05-01

    A new radiochromic film, the yellow Gafchromic EBT2, has been marketed as a drop-in replacement for the discontinued blue EBT film. In order to verify the manufacturer's claims prior to clinical use, EBT2 was characterized in transmission, and the less commonly used, reflection modes with an Epson Expression 10000XL A3 flatbed scanner. The red channel was confirmed to provide the greatest sensitivity and was used for all measurements. The post-irradiation darkening of the film was investigated, and the relative response was found to be dose dependent with higher doses stabilizing earlier than lower doses. After 13 h all dose levels had stabilized to within 1% of their value at 24 h. Uniformity of irradiated EBT2 films was within 0.8% and 1.2% (2SD of signal), in reflection and transmission modes, respectively. The light scattering effect, arising from the structure and thickness of EBT2, was found to give rise to an apparent scanner non-uniformity of up to 5.5% in signal. In reflection mode, differences of up to 1.2% were found between the signal obtained from a small film fragment (5 × 5 cm2) and the signal obtained from the same fragment bordered by extra film. Further work is needed to determine the origin of this effect, as there will be implications for reflection dosimetry of intensity modulated fields; reflection mode cannot yet be regarded as a viable alternative to transmission mode. Our results suggest that EBT2 film is a valid alternative, rather than a direct replacement for EBT film.

  20. SU-E-T-445: Lateral Optical Density Variation in Flatbed Scanners in Combination with Gafchromic Film

    SciTech Connect

    Battum, LJ van; Heukelom, S

    2014-06-01

    Purpose This study investigates the origin of lateral optical density (OD) variation for Gafchromic film (EBT and EBT2) scanned in transmission mode with Epson flatbed scanners (1680 Expression Pro and 10000XL). Effects investigated are: cross talk, optical path length and polarization. Methods Cross talk has been examined with triangular shaped light-transmission sheets with OD ranging from 0 to opaque. Optical path length has been studied with absorptive and reflective OD-filters (OD range 0.2 to 2.0). Dependency on light-polarization on the scanner read out has been investigated using linear polarizer sheets. All experiments have been performed at centre scanner position (norm point) and at several lateral scan positions, without and with (un)irradiated EBT-film. Dose values used ranged between 0.2 to 9 Gy, yielding an OD-range between 0.25 to 1.1. Results The lateral OD variation is dose dependent and increases up to 14% at most lateral position for dose up to 9 Gy. Cross talk effect contributes to 0.5% in clinical used OD ranges but equals 2% for extreme high dose gradients. Film induced optical path length will effect the lateral OD variation up to 3% at most lateral points. Light polarization is inherent present in these scanners due to multiple reflection on mirrors. In addition film induced polarization is the most important effect generating the observed lateral OD variation. Both Gafchromic film base and sensitive layer have polarizing capabilities; for the sensitive layer its influence is dose dependent. Conclusions Lateral OD variation origins from optical physics (i.e. polarization and reflection) related to scanner and film construction. Cross talk can be ignored in film dosimetry for clinical used dose values and gradients. Therefore it is recommended to determine the lateral OD variation per film type and scanner.

  1. Verification of Gamma Knife extend system based fractionated treatment planning using EBT2 film

    SciTech Connect

    Natanasabapathi, Gopishankar; Bisht, Raj Kishor

    2013-12-15

    Purpose: This paper presents EBT2 film verification of fractionated treatment planning with the Gamma Knife (GK) extend system, a relocatable frame system for multiple-fraction or serial multiple-session radiosurgery.Methods: A human head shaped phantom simulated the verification process for fractionated Gamma Knife treatment. Phantom preparation for Extend Frame based treatment planning involved creating a dental impression, fitting the phantom to the frame system, and acquiring a stereotactic computed tomography (CT) scan. A CT scan (Siemens, Emotion 6) of the phantom was obtained with following parameters: Tube voltage—110 kV, tube current—280 mA, pixel size—0.5 × 0.5 and 1 mm slice thickness. A treatment plan with two 8 mm collimator shots and three sectors blocking in each shot was made. Dose prescription of 4 Gy at 100% was delivered for the first fraction out of the two fractions planned. Gafchromic EBT2 film (ISP Wayne, NJ) was used as 2D verification dosimeter in this process. Films were cut and placed inside the film insert of the phantom for treatment dose delivery. Meanwhile a set of films from the same batch were exposed from 0 to 12 Gy doses for calibration purposes. An EPSON (Expression 10000 XL) scanner was used for scanning the exposed films in transparency mode. Scanned films were analyzed with inhouse written MATLAB codes.Results: Gamma index analysis of film measurement in comparison with TPS calculated dose resulted in high pass rates >90% for tolerance criteria of 1%/1 mm. The isodose overlay and linear dose profiles of film measured and computed dose distribution on sagittal and coronal plane were in close agreement.Conclusions: Through this study, the authors propose treatment verification QA method for Extend frame based fractionated Gamma Knife radiosurgery using EBT2 film.

  2. EBT2 film as a depth-dose measurement tool for radiotherapy beams over a wide range of energies and modalities

    SciTech Connect

    Arjomandy, Bijan; Tailor, Ramesh; Zhao Li; Devic, Slobodan

    2012-02-15

    Purpose: One of the fundamental parameters used for dose calculation is percentage depth-dose, generally measured employing ionization chambers. There are situations where use of ion chambers for measuring depth-doses is difficult or problematic. In such cases, radiochromic film might be an alternative. The EBT-2 model GAFCHROMIC film was investigated as a potential tool for depth-dose measurement in radiotherapy beams over a broad range of energies and modalities. Methods: Pieces of the EBT-2 model GAFCHROMIC EBT2 film were exposed to x-ray, electron, and proton beams used in radiotherapy. The beams employed for this study included kilovoltage x-rays (75 kVp), {sup 60}Co gamma-rays, megavoltage x-rays (18 MV), electrons (7 and 20 MeV), and pristine Bragg-peak proton beams (126 and 152 MeV). At each beam quality, film response was measured over the dose range of 0.4-8.0 Gy, which corresponds to optical densities ranging from 0.05 to 0.4 measured with a flat-bed document scanner. To assess precision in depth-dose measurements with the EBT-2 model GAFCHROMIC film, uncertainty in measured optical density was investigated with respect to variation in film-to-film and scanner-bed uniformity. Results: For most beams, percentage depth-doses measured with the EBT-2 model GAFCHROMIC film show an excellent agreement with those measured with ion chambers. Some discrepancies are observed in case of (i) kilovoltage x-rays at larger depths due to beam-hardening, and (ii) proton beams around Bragg-peak due to quenching effects. For these beams, an empirical polynomial correction produces better agreement with ion-chamber data. Conclusions: The EBT-2 model GAFCHROMIC film is an excellent secondary dosimeter for measurement of percentage depth-doses for a broad range of beam qualities and modalities used in radiotherapy. It offers an easy and efficient way to measure beam depth-dose data with a high spatial resolution.

  3. Note: Calibration of EBT3 radiochromic film for measuring solar ultraviolet radiation

    SciTech Connect

    Chun, S. L.; Yu, P. K. N.

    2014-10-01

    Solar (UVA + UVB) exposure was assessed using the Gafchromic EBT3 film. The coloration change was represented by the net reflective optical density (Net ROD). Through calibrations against a UV-tube lamp, operational relationships were obtained between Net ROD and the (UVA + UVB) exposures (in J cm⁻²p or J m⁻²). The useful range was from ~0.2 to ~30 J cm⁻². The uniformity of UV irradiation was crucial for an accurate calibration. For solar exposures ranging from 2 to 11 J cm⁻², the predicted Net ROD agreed with the recorded values within 9%, while the predicted exposures agreed with the recorded values within 15%.

  4. Companding technique for high dynamic range measurements using Gafchromic films

    SciTech Connect

    Van den Heuvel, Frank; Crijns, Wouter; Defraene, Gilles

    2011-12-15

    Purpose: To introduce a methodology to perform dose measurements using Gafchromic films which can span several decades of dose levels. Methods: The technique is based on a rescaling approach using different films irradiated at different dose levels. This is combined with a registration protocol correcting positioning and scaling factors for each film. The methodology is validated using TLD's for out-of-field doses. Furthermore, two examples are provided using the technique to characterize small sized radiosurgery cones and compared with measurements made with a pinpoint chamber. Results: Excellent agreement with TLD, planning systems and measurement was found. The superior resolution of the film technique was apparent. Conclusions: The authors have introduced a new technique allowing users to quantify very low doses in conjunction with commissioning measurements. The use of film also provides 2D information on beam characteristics in high resolution measurements.

  5. Changes of the optical characteristics of radiochromic films in the transition from EBT3 to EBT-XD films

    NASA Astrophysics Data System (ADS)

    Schoenfeld, Andreas A.; Wieker, Soeren; Harder, Dietrich; Poppe, Bjoern

    2016-07-01

    A new type of radiochromic film, the EBT-XD film, has been introduced with the aim to reduce the orientation effect and the lateral response artifact occurring in the use of radiochromic films together with flatbed scanners. The task of the present study is to quantify the changes of optical characteristics involved with the transition from the well-known EBT3 films to the new EBT-XD films, using the optical bench arrangement already applied by Schoenfeld et al (2014 Phys. Med. Biol. 59 3575–97). Largely reduced polarization effects and the almost complete loss of the anisotropy of the scattered light produced in a radiation-exposed film have been observed. The Rayleigh–Debye–Gans theory is used to understand these optical changes as arising from the reduced length-to-width ratio of the LiPCDA polymer crystals in the active layer of the EBT-XD film. The effect of these changes on the flatbed scanning artifacts will be shortly addressed, but treated in more detail in a further paper.

  6. Outdoor solar UVA dose assessment with EBT2 radiochromic film using spectrophotometer and densitometer measurements.

    PubMed

    Abukassem, I; Bero, M A

    2015-04-01

    Direct measurements of solar ultraviolet radiations (UVRs) have an important role in the protection of humans against UVR hazard. This work presents simple technique based on the application of EBT2 GAFCHROMIC(®) film for direct solar UVA dose assessment. It demonstrates the effects of different parts of the solar spectrum (UVB, visible and infrared) on performed UVA field measurements and presents the measurement uncertainty budget. The gradient of sunlight exposure level permitted the authors to establish the mathematical relationships between the measured solar UVA dose and two measured quantities: the first was the changes in spectral absorbance at the wavelength 633 nm (A633) and the second was the optical density (OD). The established standard relations were also applied to calculate the solar UVA dose variations during the whole day; 15 min of exposure each hour between 8:00 and 17:00 was recorded. Results show that both applied experimental methods, spectrophotometer absorbance and densitometer OD, deliver comparable figures for EBT2 solar UVA dose assessment with relative uncertainty of 11% for spectral absorbance measurements and 15% for OD measurements. PMID:25500756

  7. SU-E-T-118: Dose Verification for Accuboost Applicators Using TLD, Ion Chamber and Gafchromic Film Measurements

    SciTech Connect

    Chisela, W; Yao, R; Dorbu, G

    2014-06-01

    Purpose: To verify dose delivered with HDR Accuboost applicators using TLD, ion chamber and Gafchromic film measurements and to examine applicator leakage. Methods: A microSelectron HDR unit was used to deliver a dose of 50cGy to the mid-plane of a 62mm thick solid water phantom using dwell times from Monte Carlo pre-calculated nomograms for a 60mm, 70mm Round and 60mm Skin-Dose Optimized (SDO) applicators respectively. GafChromic EBT3+ film was embedded in the phantom midplane horizontally to measure dose distribution. Absolute dose was also measured with TLDs and an ADCL calibrated parallel-plate ion chamber placed in the film plane at field center for each applicator. The film was calibrated using 6MV x-ray beam. TLDs were calibrated in a Cs-137 source at UW-Madison calibration laboratory. Radiation leakage through the tungsten alloy shell was measured with a film wrapped around outside surface of a 60mm Round applicator. Results: Measured maximum doses at field center are consistently lower than predicated by 5.8% for TLD, 8.8% for ion chamber, and 2.6% for EBT3+ film on average, with measurement uncertainties of 2.2%, 0.3%, and 2.9% for TLD, chamber, film respectively. The total standard uncertainties for ion chamber and Gafchromic film measurement are 4.9% and 4.6% respectively[1]. The area defined by the applicator aperture was covered by 80% of maximum dose for 62mm compression thickness. When 100cGy is delivered to mid-plane with a 60mm Round applicator, surface dose ranges from 60cGy to a maximum of 145cGy, which occurs at source entrance to the applicator. Conclusion: Measured doses by all three techniques are consistently lower than predicted in our measurements. For a compression thickness of 62 mm, the field size defined by the applicator is only covered by 80% of prescribed dose. Radiation leakage of up to 145cGy was found at the source entrance of applicators.

  8. A protocol for EBT3 radiochromic film dosimetry using reflection scanning

    SciTech Connect

    Papaconstadopoulos, Pavlos Hegyi, Gyorgy; Seuntjens, Jan; Devic, Slobodan

    2014-12-15

    Purpose: To evaluate the performance of the EBT3 radiochromic film dosimetry system using reflection measurements and to suggest a calibration protocol for precise and accurate reflection film dosimetry. Methods: A set of 14 Gafchromic EBT3 film pieces were irradiated to various doses ranging from 0 to 8 Gy and subsequently scanned using both the reflection and transmission mode. Scanning resolution varied from 50 to 508 dpi (0.5–0.05 mm/pixel). Both the red and green color channels of scanned images were used to relate the film response to the dose. A sensitivity, uncertainty, and accuracy analysis was performed for all scanning modes and color channels. The total uncertainty, along with the fitting and experimental uncertainty components, was identified and analyzed. A microscope resolution target was used to evaluate possible resolution losses under reflection scanning. The calibration range was optimized for reflection scanning in the low (<2 Gy) and high (>2 Gy) dose regions based on the reported results. Results: Reflection scanning using the red channel exhibited the highest sensitivity among all modes, being up to 150% higher than transmission mode in the red channel for the lowest dose level. Furthermore, there was no apparent loss in resolution between the two modes. However, higher uncertainties and reduced accuracy were observed for the red channel under reflection mode, especially at dose levels higher than 2 Gy. These uncertainties were mainly attributed to saturation effects which were translated in poor fitting results. By restricting the calibration to the 0–2 Gy dose range, the situation is reversed and the red reflection mode was superior to the transmission mode. For higher doses, the green channel in reflection mode presented comparable results to the red transmission. Conclusions: A two-color reflection scanning protocol can be suggested for EBT3 radiochromic film dosimetry using the red channel for doses less than 2 Gy and the green

  9. Comparison of the Epson Expression 1680 flatbed and the Vidar VXR-16 Dosimetry PRO trade mark sign film scanners for use in IMRT dosimetry using Gafchromic and radiographic film

    SciTech Connect

    Wilcox, Ellen; Daskalov, George; Nedialkova, Lucy

    2007-01-15

    Intensity-modulated radiotherapy (IMRT) treatment plan verification is often done using Kodak EDR2 film and a Vidar Dosimetry PRO trade mark sign film digitizer. However, since many hospitals are moving towards a filmless environment, access to a film processor may not be available. Therefore, we have investigated a newly available Gafchromic[reg] EBT film for IMRT dosimetry. Planar IMRT dose distributions are delivered to both EBT and EDR2 film and scanned with the Vidar VXR-16 as well as an Epson Expression 1680 flatbed scanner. The measured dose distributions are then compared to those calculated with a Pinnacle treatment planning system. The IMRT treatments consisted of 7-9 6 MV beams for treatment of prostate, head and neck, and a few other sites. The films were analyzed using FilmQA trade mark sign (3cognition LLC) software. Comparisons between measured and calculated dose distributions are reported as dose difference (DD) (pixels within {+-}5%), distance to agreement (DTA) (3 mm), as well as gamma values ({gamma}) (dose={+-}3%, dist.=2 mm). Using EDR2 with the Vidar scanner is an established technique and agreement between calculated and measured dose distributions was better than 90% in all indices (DD, DTA, and {gamma}). However, agreement with calculations deteriorated reaching the lower 80% for EBT film scans with the Vidar scanner in logarithmic mode. The EBT Vidar scans obtained in linear mode showed an improved agreement to the upper 80% range, but artifacts were still observed across the scan. These artifacts were very distinct in all EBT scans and can be attributed to the way the film is transported through the scanner. In the Epson scanner both films are rigidly immobilized and the light source scans over the film. It was found that the Epson scanner performed equally well with both types of film giving agreement to better than 90% in all indices.

  10. Application of a color scanner for 60Co high dose rate brachytherapy dosimetry with EBT radiochromic film

    PubMed Central

    Ghorbani, Mahdi; Toossi, Mohammad Taghi Bahreyni; Mowlavi, Ali Asghar; Roodi, Shahram Bayani; Meigooni, Ali Soleimani

    2012-01-01

    Background. The aim of this study is to evaluate the performance of a color scanner as a radiochromic film reader in two dimensional dosimetry around a high dose rate brachytherapy source. Materials and methods A Microtek ScanMaker 1000XL film scanner was utilized for the measurement of dose distribution around a high dose rate GZP6 60Co brachytherapy source with GafChromic® EBT radiochromic films. In these investigations, the non-uniformity of the film and scanner response, combined, as well as the films sensitivity to scanner’s light source was evaluated using multiple samples of films, prior to the source dosimetry. The results of these measurements were compared with the Monte Carlo simulated data using MCNPX code. In addition, isodose curves acquired by radiochromic films and Monte Carlo simulation were compared with those provided by the GZP6 treatment planning system. Results Scanning of samples of uniformly irradiated films demonstrated approximately 2.85% and 4.97% nonuniformity of the response, respectively in the longitudinal and transverse directions of the film. Our findings have also indicated that the film response is not affected by the exposure to the scanner’s light source, particularly in multiple scanning of film. The results of radiochromic film measurements are in good agreement with the Monte Carlo calculations (4%) and the corresponding dose values presented by the GZP6 treatment planning system (5%). Conclusions The results of these investigations indicate that the Microtek ScanMaker 1000XL color scanner in conjunction with GafChromic EBT film is a reliable system for dosimetric evaluation of a high dose rate brachytherapy source. PMID:23411947

  11. Poster — Thur Eve — 53: Novel Technique for the Measurement of Ultra-Superficial Doses Using Gafchromic Film

    SciTech Connect

    Marcos, M.; Devic, S.

    2014-08-15

    Purpose: Dose build-up and electron contamination are two closely related quantities with important implications in radiotherapy, yet they are quite difficult to measure with great certainty. We present a novel technique for measuring ultra-superficial doses. Method and Materials: We used Gafchromic EBT-3 film which have an effective point of measurement of roughly 153 micros (effective depth in water). By peeling off one of the polyester layers, the active layer becomes the top layer and we obtain a film with a effective point of measurement of 15 microns (effective depth in water). A film calibration was performed using a 180 kVp orthovoltage beam. Since the active layer of the film may have been compressed or perturbed during the peeling of clear polyester we use a triple-channel film calibration technique to minimize the effects of non-uniformity in the active layer. We measured surface doses of orthovoltage beams with lead cutouts in place to introduce contaminant photoelectrons. Results: Our measurements show that the dose enhancement near the edges of the lead were about 125% relative to central axis for 6 cm diameter cutouts up to 170% for 2 cm diameter cutouts, which were within 5% of our EGSnrc based Monte Carlo simulations.

  12. Direct measurement of absorbed dose to water in HDR {sup 192}Ir brachytherapy: Water calorimetry, ionization chamber, Gafchromic film, and TG-43

    SciTech Connect

    Sarfehnia, Arman; Kawrakow, Iwan; Seuntjens, Jan

    2010-04-15

    Purpose: Gafchromic film and ionometric calibration procedures for HDR {sup 192}Ir brachytherapy sources in terms of dose rate to water are presented and the experimental results are compared to the TG-43 protocol as well as with the absolute dose measurement results from a water calorimetry-based primary standard. Methods: EBT-1 Gafchromic films, an A1SL Exradin miniature Shonka thimble type chamber, and an SI HDR 1000 Plus well-type chamber (Standard Imaging, Inc., Middleton, WI) with an ADCL traceable S{sub k} calibration coefficient (following the AAPM TG-43 protocol) were used. The Farmer chamber and Gafchromic film measurements were performed directly in water. All results were compared to direct and absolute absorbed dose to water measurements from a 4 deg. C stagnant water calorimeter. Results: Based on water calorimetry, the authors measured the dose rate to water to be 361{+-}7 {mu}Gy/(h U) at a 55 mm source-to-detector separation. The dose rate normalized to air-kerma strength for all the techniques agree with the water calorimetry results to within 0.83%. The overall 1-sigma uncertainty on water calorimetry, ionization chamber, Gafchromic film, and TG-43 dose rate measurement amounts to 1.90%, 1.44%, 1.78%, and 2.50%, respectively. Conclusions: This work allows us to build a more realistic uncertainty estimate for absorbed dose to water determination using the TG-43 protocol. Furthermore, it provides the framework necessary for a shift from indirect HDR {sup 192}Ir brachytherapy dosimetry to a more accurate, direct, and absolute measurement of absorbed dose to water.

  13. Validation of GEANT4 simulations for percentage depth dose calculations in heterogeneous media by using small photon beams from the 6-MV Cyberknife: Comparison with photon beam dosimetry with EBT2 film

    NASA Astrophysics Data System (ADS)

    Lee, Chung Il; Yoon, Sei-Chul; Shin, Jae Won; Hong, Seung-Woo; Suh, Tae Suk; Min, Kyung Joo; Lee, Sang Deok; Chung, Su Mi; Jung, Jae-Yong

    2015-04-01

    Percentage depth dose (PDD) distributions in heterogeneous phantoms with lung and soft bone equivalent media are studied by using the GEANT4 Monte Carlo code. For lung equivalent media, Balsa wood is used, and for soft bone equivalent media, a compound material with epoxy resin, hardener and calcium carbonate is used. Polystyrene slabs put together with these materials are used as a heterogeneous phantom. Dose measurements are performed with Gafchromic EBT2 film by using photon beams from the 6-MV CyberKnife at the Seoul Uridul Hospital. The cone sizes of the photon beams are varied from 5 to 10 to 30 mm. When the Balsa wood is inserted in the phantom, the dose measured with EBT2 film is found to be significantly different from the dose without the EBT2 film in and the dose beyond the Balsa wood region, particularly for small field sizes. On the other hand, when the soft bone equivalent material is inserted in the phantom, the discrepancy between the dose measured with EBT2 film and the dose without EBT2 film can be seen only in the region of the soft bone equivalent material. GEANT4 simulations are done with and without the EBT2 film to compare the simulation results with measurements. The GEANT4 simulations including EBT2 film are found to agree well with the measurements for all the cases within an error of 2.2%. The results of the present study show that GEANT4 gives reasonable results for the PDD calculations in heterogeneous media when using photon beams produced by the 6-MV CyberKnife

  14. Dosimetric properties of improved GafChromic films for seven different digitizers.

    PubMed

    Devic, Slobodan; Seuntjens, Jan; Hegyi, Gyorgy; Podgorsak, Ervin B; Soares, Christopher G; Kirov, Assen S; Ali, Imad; Williamson, Jeffrey F; Elizondo, Angel

    2004-09-01

    Two recently introduced GafChromic film models, HS and XR-T, have been developed as more sensitive and uniform alternatives to GafChromic MD-55-2 film. The HS model has been specifically designed for measurement of absorbed dose in high-energy photon beams (above 1 MeV), while the XR-T model has been introduced for dose measurements of low energy (0.1 MeV) photons. The goal of this study is to compare the sensitometric curves and estimated dosimetric uncertainties associated with seven different GafChromic film dosimetry systems for the two new film models. The densitometers tested are: LKB Pharmacia UltroScan XL, Molecular Dynamics Personal Densitometer, Nuclear Associates Radiochromic Densitometer Model 37-443, Photoelectron Corporation CMR-604, Laser Pro 16, Vidar VXR-16, and AGFA Arcus II document scanner. Pieces of film were exposed to different doses in a dose range from 0.5 to 50 Gy using 6 MV photon beam. Functional forms for dose vs net optical density have been determined for each of the GafChromic film-dosimetry systems used in this comparison. Two sources of uncertainties in dose measurements, governed by the experimental measurement and calibration curve fit procedure, have been compared for the densitometers used. Among the densitometers tested, it is found that for the HS film type the uncertainty caused by the experimental measurement varies from 1% to 3% while the calibration fit uncertainty ranges from 2% to 4% for doses above 5 Gy. Corresponding uncertainties for XR-T film model are somewhat higher and range from 1% to 5% for experimental and from 2% to 7% for the fit uncertainty estimates. Notwithstanding the significant variations in sensitivity, the studied densitometers exhibit very similar precision for GafChromic film based dose measurements above 5 Gy. PMID:15487718

  15. SU-E-T-12: A Feasibility Study of Patient Specific QA Using Gafchromic Film of Dynamic Feathering in Junctions of Craniospinal Irradiation

    SciTech Connect

    Stanford, J; Duggar, W; Yang, C

    2014-06-01

    Purpose: Cranio-spinal irradiation is the most complicated format of the conventional external beam radiation therapy because it involves matches of non-coplanar beams which are susceptible to daily setup errors. This study explores the efficacy of Gafchromic film dosimetry to quantitatively verify the junctions for cranio-spinal radiation feathered with field in field technique. Methods: 15cm in thickness of solid water phantom was scanned vertically and exported to the Pinnacle TPS as primary phantom data set. A patient cranio-spinal plan, consisted of two bilateral whole brain beams dynamically matched with a posterior spinal beam using field in field technique, was transferred to the phantom and recalculated for one fraction with set monitor units identical to the original plan. Next, planar dose distribution on the phantom was exported to the FilmQA Pro 2013 software (Ashland, Inc.) in binary format for comparison with the measured dose distribution. An EBT2 film was sandwiched in the middle of the phantom and the phantom was set up according to the QA plan based on the room laser system. The shifts instructions associated with the patient original plan were made and the beams from the patient original plan delivered to the solid water phantom via the record and verify system in QA mode. The dose distribution from the measured film was compared with the planned reference distribution using gamma analysis and profile comparison. Results: Gamma passing rate of 91 % with DTA 3mm and 5% dose difference was obtained within the junction region, significantly greater passing rate above 95 % was obtained in the homogeneous region of the brain field. Conclusion: This study confirms that Gafchromic film dosimetry can be used to validate the efficacy of FIF feathering technique for cranio-spinal treatment. FIF technique with Gafchromic dosimetry may now be the new standard for delivering efficient and accurate cranio-spinal radiation with confidence.

  16. Dose area product evaluations with Gafchromic XR-R films and a flat-bed scanner.

    PubMed

    Rampado, O; Garelli, E; Deagostini, S; Ropolo, R

    2006-12-01

    Gafchromic XR-R films are a useful tool to evaluate entrance skin dose in interventional radiology. Another dosimetric quantity of interest in diagnostic and interventional radiology is the dose area product (DAP). In this study, a method to evaluate DAP using Gafchromic XR-R films and a flat-bed scanner was developed and tested. Film samples were exposed to an x-ray beam of 80 kVp over a dose range of 0-10 Gy. DAP measurements with films were obtained from the digitalization of a film sample positioned over the x-ray beam window during the exposure. DAP values obtained with this method were compared for 23 cardiological interventional procedures with DAP values displayed by the equipment. The overall one-sigma dose measurement uncertainty depended on the absorbed dose, with values below 6% for doses above 1 Gy. A maximum discrepancy of 16% was found, which is of the order of the differences in the DAP measurements that may occur with different calibration procedures. Based on the results presented, after an accurate calibration procedure and a thorough inspection of the relationship between the actual dose and the direct measured quantity (net optical density or net pixel value variation), Gafchromic XR-R films can be used to assess the DAP. PMID:17110757

  17. Characterization of differences in calculated and actual measured skin doses to canine limbs during stereotactic radiosurgery using Gafchromic film

    SciTech Connect

    Walters, Jerri; Ryan, Stewart; Harmon, Joseph F.

    2012-07-01

    Accurate calculation of absorbed dose to the skin, especially the superficial and radiosensitive basal cell layer, is difficult for many reasons including, but not limited to, the build-up effect of megavoltage photons, tangential beam effects, mixed energy scatter from support devices, and dose interpolation caused by a finite resolution calculation matrix. Stereotactic body radiotherapy (SBRT) has been developed as an alternative limb salvage treatment option at Colorado State University Veterinary Teaching Hospital for dogs with extremity bone tumors. Optimal dose delivery to the tumor during SBRT treatment can be limited by uncertainty in skin dose calculation. The aim of this study was to characterize the difference between measured and calculated radiation dose by the Varian Eclipse (Varian Medical Systems, Palo Alto, CA) AAA treatment planning algorithm (for 1-mm, 2-mm, and 5-mm calculation voxel dimensions) as a function of distance from the skin surface. The study used Gafchromic EBT film (International Specialty Products, Wayne, NJ), FilmQA analysis software, a limb phantom constructed from plastic water Trade-Mark-Sign (fluke Biomedical, Everett, WA) and a canine cadaver forelimb. The limb phantom was exposed to 6-MV treatments consisting of a single-beam, a pair of parallel opposed beams, and a 7-beam coplanar treatment plan. The canine forelimb was exposed to the 7-beam coplanar plan. Radiation dose to the forelimb skin at the surface and at depths of 1.65 mm and 1.35 mm below the skin surface were also measured with the Gafchromic film. The calculation algorithm estimated the dose well at depths beyond buildup for all calculation voxel sizes. The calculation algorithm underestimated the dose in portions of the buildup region of tissue for all comparisons, with the most significant differences observed in the 5-mm calculation voxel and the least difference in the 1-mm voxel. Results indicate a significant difference between measured and calculated data

  18. SU-D-304-03: Small Field Proton Dosimetry Using MicroDiamond and Gafchromic Film

    SciTech Connect

    Andersen, A; Das, I; Coutinho, L

    2015-06-15

    Purpose: Certain dosimetric characteristics continue to make proton beam therapy an appealing modality for cancer treatment. The proton Bragg peak allows for conformal radiation dose delivery to the target while reducing dose to normal tissue and organs. As field sizes become very small the benefit of the Bragg peak is diminished due to loss of transverse equilibrium along the central beam axis. Furthermore, aperture scattering contributes additional dose along the central axis. These factors warrant the need for accurate small field dosimetry. In this study small field dosimetry was performed using two different methods. Methods: Small field dosimetry measurements were performed using a PTW microdiamond detector and Gafchromic EBT2 film for aperture sizes ranging from 0.5cm to 10cm and a proton range in water of 10cm to 27cm. The measurements were analyzed and then compared to each other and to reference dosimetry data acquired with a Markus chamber. Results: A decrease in normalized output is observed at small field sizes and at larger ranges in water using both measurement methods. Also, a large variation is observed between the output measurements by microdiamond and film at very small field sizes. At the smallest aperture, normalized output ranged from 0.16 to 0.72 and the percent difference between both measurement methods ranged from 36% to 70% depending on proton range. At field sizes above 5cm the film and microdiamond agree within 3%. Conclusion: Although both measurement methods exhibit a general decrease in output factor at small field sizes, dosimetric measurements for small fields using these two methods can vary significantly. Dosimetry under standard conditions is not sufficient to correctly model the dose distributions and outputs factors for small field sizes, additional small field measurements should be performed.

  19. The Use of TLD and Gafchromic Film to Assure Submillimeter Accuracy for Image-Guided Radiosurgery

    SciTech Connect

    Ho, Anthony K. Gibbs, Iris C.; Chang, Steve D.; Main, Bill; Adler, John R.

    2008-04-01

    The Cyberknife is an image-guided radiosurgical system. It uses a compact X-band 6-MV linear accelerator mounted on a robotic arm to deliver radiosurgical doses. While routine quality assurance (QA) is essential for any radiosurgery system, QA plays an even more vital role for the Cyberknife system, due to the complexity of the system and the wide range of applications. This paper presents a technique for performing quality assurance using thermoluminescence detectors (TLDs) and Gafchromic films that is intended to be specific for the Cyberknife. However, with minor modification, the proposed method can also be used for QA of other radiosurgery systems. Our initial QA procedure for the CyberKnife utilized a 30 x 30 x 11-cm solid water phantom containing a planar array of slots for 1x 1 x 1-mm TLDs on a 2-mm grid. With the objective of significantly simplifying CyberKnife QA, a new procedure for verification was developed, which uses much fewer TLDs than the prior solid water phantom technique. This new method requires only that the system target dose to the center of a cluster of 7 TLDs. In a prior study with Gafchromic films, conducted at 3 different Cyberknife facilities, the mean clinically relevant error was demonstrated to be 0.7 mm. A similar Gafchromic film analysis replicated these error measurements as part of the present investigation. It cannot be emphasized enough the importance of implementing routine QA to verify the accuracy of any radiosurgery system. Our quality assurance procedure tests the treatment planning system, as well as the entire treatment delivery including the image targeting system and the robot system. Either TLDs or Gafchromic films may be used for QA test of a radiosurgery system. Using both methods for measurement has the advantage independently verifying the accuracy of the system. This approach, which is routinely in used at our institution, has repeatedly confirmed the submillimeter targeting accuracy of our Cyberknife.

  20. Practical IMRT QA dosimetry using Gafchromic film: a quick start guide.

    PubMed

    Bennie, Nick; Metcalfe, Peter

    2016-06-01

    This work outlines a method for using Gafchromic film for dosimetry purposes, by scanning it with currently available commercial scanners. The scanners used were: Epson V800, Epson V700, Epson V37 series, specifically a V370 and a Canon multi-function office printer/scanner. The Epson scanners have 16 bit RGB resolution, the Canon has 8 bit RGB (Red Green Blue) resolution, and the V800 and V700 allow scanning in transmission mode. The V700 uses an Epson White Cold Cathode Florescent Lamp; the recently released V800 uses an Epson light emitting diode (LED) light source, while the V37 series uses a reflective mode and the Epson LED light source. The Epson V37 series scanners are designed for non-professional use so the cost has been kept at a low "entry level" point, so they would be a suitable option for a department wanting to use Gafchromic film or with limited needs that did not justify a more sophisticated and expensive unit. Note that the V800 or V700 scanners are not expensive in context, costing approximately the same as a 25 sheet box of Gafchromic film. The Canon was included to demonstrate that a scanner with 8 bit RGB resolution can be used for dosimetry. These general multi-function units are available in most departments, and they would allow Gafchromic film to be evaluated as a dosimetry tool without a significant investment. Furthermore, they are generally capable of scanning large format film (425 × 350 mm) in one part. Although this is not necessary for dosimetry, it is often useful for machine QA, where dividing the film into two parts to ensure accurate measurements is not practical. Moreover, this analytical method uses software that is freely or commonly available, particularly the image processing package ImageJ. Note ImageJ v1.48 was the version used. The results demonstrate that this method used with the scanners evaluated is a practical method of using Gafchromic film as a dosimeter for IMRT QA. PMID:27098156

  1. Thermal neutron imaging through XRQA2 GAFCHROMIC films coupled with a cadmium radiator

    NASA Astrophysics Data System (ADS)

    Sacco, D.; Bedogni, R.; Bortot, D.; Palomba, M.; Pola, A.; Introini, M. V.; Lorenzoli, M.; Gentile, A.; Strigari, L.; Pressello, C.; Soriani, A.; Gómez-Ros, J. M.

    2015-10-01

    A simple and inexpensive method to perform passive thermal neutron imaging on large areas was developed on the basis of XRQA2 GAFCHROMIC films, commonly employed for quality assurance in radiology. To enhance their thermal neutron response, the sensitive face of film was coupled with a 1 mm thick cadmium radiator, forming a sandwich. By exchanging the order of Cd filter and sensitive film with respect to the incident neutron beam direction, two different configurations (beam-Cd-film and beam-film-Cd) were identified. These configurations were tested at thermal neutrons fluence values in the range 109-1010 cm-2, using the ex-core radial thermal neutron column of the ENEA Casaccia - TRIGA reactor. The results are presented in this work.

  2. Use of GafChromic film to diagnose laser generated proton beams

    SciTech Connect

    Hey, D. S.; Key, M. H.; Mackinnon, A. J.; MacPhee, A. G.; Patel, P. K.; Freeman, R. R.; Van Woerkom, L. D.; Castaneda, C. M.

    2008-05-15

    A calibration of three types of GafChromic radiochromic film (HS, MD-55, and HD-810) was carried out on the Crocker Nuclear Laboratory's 76 in. cyclotron at UC Davis over doses ranging from 0.001 to 15 kGy. The film was digitized with a scanning microdensitometer with which it was scanned twice with two different filters to increase the film's effective dynamic range. We demonstrate how this calibrated film can be used to measure the spectrum and total energy of a laser generated proton beam. This technique was applied to an experiment on the 10 J, 100 fs Callisto laser at Lawrence Livermore National Laboratory. The resulting proton spectrum was compared to that obtained by simultaneous measurement of Ti nuclear activation; the two methods give the same proton beam slope temperature and agree in number of protons to within 27%.

  3. An application of GafChromic MD-55 film for 67.5 MeV clinical proton beam dosimetry.

    PubMed

    Daftari, I; Castenadas, C; Petti, P L; Singh, R P; Verhey, L J

    1999-11-01

    The purpose of this study is to explore the use of GafChromic MD-55 (RC) film for 67.5 MeV clinical proton beam dosimetry at the Crocker Nuclear Laboratory, University of California, Davis. Several strips of RC film 6 cm x 6 cm in dimension were irradiated at a depth of 18.2 mm corresponding to the middle of a 24 mm spread-out Bragg peak (SOBP). The films were irradiated to a proton dose in the range of 0.5 Gy to 100 Gy. The beam profiles were also measured at the middle of the 24 mm SOBP. The Bragg peak was measured by using a wedge shaped phantom made of Lucite. The Bragg peak measured with RC film was compared with diode and ionization chamber measurements. After background subtraction, the calibration of the dose response of RC film showed, to a maximum deviation of 10%, a linear increase of optical density (OD) with dose from 0.5 to 100 Gy. The uniformity of OD over a single sheet of film showed a variation of +/-6%. The distal-fall off between 90% and 20% measured with GafChromic film for the Bragg peak was 1.3 mm as compared to 1.1 mm for a diode measurement and 1.4 mm for an ionization chamber measurement. The FWHM of the Bragg peak was 7.5 mm when measured with GafChromic film, 5.3 mm when measured with a diode and 8.1 mm as measured by an ionization chamber. The peak/plateau ratio with GafChromic film was 3.3 as compared to 3.7 with a diode and 3.2 with an ionization chamber. In conclusion, GafChromic MD-55 film may be a useful and convenient detector for dose measurement and quality assurance programmes of proton beams. PMID:10588281

  4. Extension of Radiochromic Film Dosimetry to Three Dimensions: A Feasibility Study

    SciTech Connect

    Adamczyk, S.; Skorska, M.

    2007-11-26

    The feasibility of measuring dose distributions in three-dimensions using Gafchromic EBT film was studied. A stack of film was used as tomographic dosimeter. The response of film (4x4 cm{sup 2}) placed between a stack of Gafchromic EBT films and slabs of polystyrene phantom was tested. Differences in film response for two different photon beam qualities, 6 MV and 25 MV, were observed. Measurements with 6 MV beam quality revealed that for different field sizes and depths there were no changes in response of EBT Gafchromic film when placed between polystyrene plates or sandwiched between other films. However, in the case of 25 MV beam quality, a statistically significant over-response was found for film placed in stack. We conclude that stacked-film dosimetry is feasible at 6 MV. For 25 MV photon beam quality the stack of films showed different properties than the polystyrene slabs of the same height.

  5. Dose rate dependence for different dosimeters and detectors: TLD, OSL, EBT films, and diamond detectors

    SciTech Connect

    Karsch, L.; Beyreuther, E.; Burris-Mog, T.; Kraft, S.; Richter, C.; Zeil, K.; Pawelke, J.

    2012-05-15

    Purpose: The use of laser accelerators in radiation therapy can perhaps increase the low number of proton and ion therapy facilities in some years due to the low investment costs and small size. The laser-based acceleration technology leads to a very high peak dose rate of about 10{sup 11} Gy/s. A first dosimetric task is the evaluation of dose rate dependence of clinical dosimeters and other detectors. Methods: The measurements were done at ELBE, a superconductive linear electron accelerator which generates electron pulses with 5 ps length at 20 MeV. The different dose rates are reached by adjusting the number of electrons in one beam pulse. Three clinical dosimeters (TLD, OSL, and EBT radiochromic films) were irradiated with four different dose rates and nearly the same dose. A faraday cup, an integrating current transformer, and an ionization chamber were used to control the particle flux on the dosimeters. Furthermore two diamond detectors were tested. Results: The dosimeters are dose rate independent up to 410{sup 9} Gy/s within 2% (OSL and TLD) and up to 1510{sup 9} Gy/s within 5% (EBT films). The diamond detectors show strong dose rate dependence. Conclusions: TLD, OSL dosimeters, and EBT films are suitable for pulsed beams with a very high pulse dose rate like laser accelerated particle beams.

  6. Alternate calibration method of radiochromic EBT3 film for quality assurance verification of clinical radiotherapy treatments

    NASA Astrophysics Data System (ADS)

    Park, Soah; Kang, Sei-Kwon; Cheong, Kwang-Ho; Hwang, Taejin; Yoon, Jai-Woong; Koo, Taeryool; Han, Tae Jin; Kim, Haeyoung; Lee, Me Yeon; Bae, Hoonsik; Kim, Kyoung Ju

    2016-07-01

    EBT3 film is utilized as a dosimetry quality assurance tool for the verification of clinical radiotherapy treatments. In this work, we suggest a percentage-depth-dose (PDD) calibration method that can calibrate several EBT3 film pieces together at different dose levels because photon beams provide different dose levels at different depths along the axis of the beam. We investigated the feasibility of the film PDD calibration method based on PDD data and compared the results those from the traditional film calibration method. Photon beams at 6 MV were delivered to EBT3 film pieces for both calibration methods. For the PDD-based calibration, the film pieces were placed on solid phantoms at the depth of maximum dose (dmax) and at depths of 3, 5, 8, 12, 17, and 22 cm, and a photon beam was delivered twice, at 100 cGy and 400 cGy, to extend the calibration dose range under the same conditions. Fourteen film pieces, to maintain their consistency, were irradiated at doses ranging from approximately 30 to 400 cGy for both film calibrations. The film pieces were located at the center position on the scan bed of an Epson 1680 flatbed scanner in the parallel direction. Intensity-modulated radiation therapy (IMRT) plans were created, and their dose distributions were delivered to the film. The dose distributions for the traditional method and those for the PDD-based calibration method were evaluated using a Gamma analysis. The PDD dose values using a CC13 ion chamber and those obtained by using a FC65-G Farmer chamber and measured at the depth of interest produced very similar results. With the objective test criterion of a 1% dosage agreement at 1 mm, the passing rates for the four cases of the three IMRT plans were essentially identical. The traditional and the PDD-based calibrations provided similar plan verification results. We also describe another alternative for calibrating EBT3 films, i.e., a PDD-based calibration method that provides an easy and time-saving approach

  7. Comparison of vidar dosimetry advantage pro and epson perfection V700 scanner in densitometry of radiochomic EBT2 film in measurement of high dose gradient

    NASA Astrophysics Data System (ADS)

    Bura, W.; Tangboonduangjit, P.; Damrongkijudom, N.

    2016-03-01

    Nowadays the radiochromic film is widely used to obtain dose distribution in two dimensions with high spatial resolution, less energy dependence and near tissue equivalent. It can be a commissioning tool to verify high dose gradient of dose distribution for IMRT and VMAT techniques. However, the film scanner could affect the accuracy of dose distribution if lack of precaution. In this study, the comparison between Epson perfection V700 and Vidar Dosimetry Pro Advantage (RED) is evaluated in terms of the capability to verify the 2D dose distribution for conventional and VMAT techniques. The Gafchromic® EBT2 films were read from two types of scanners (Epson perfection V700 and Vidar Dosimetry Pro Advantage) for volumetric modulated radiation therapy (VMAT) dosimetry. The software for analyzing the results of Epson perfection V700 and Vidar Dosimetry Pro Advantage are SNC Patient software and Omnipro’ IMRT software, respectively. Comparisons between measured and calculated dose distributions are reported as %passing rate and the gamma index for tolerance parameters of 3% and 3mm. The study found that the %passing rate obtained from Vidar scanner and Epson V700 scanner compared with Eclipse treatment planning system is more than 98% with the criteria of (3%/3mm).

  8. SU-E-T-176: Examination of Surface Dose Enhancement Using Radiochromic EBT3-Films in a Cylindrical Setup

    SciTech Connect

    Failing, T; Chofor, N; Poppinga, D; Schoenfeld, A; Poppe, B; Willborn, K

    2014-06-01

    Purpose: This study was undertaken to optimize the measurement techniques with radiochromic EBT3 films to offer accurate surface dose measurements and at the same time high resolution depth dose curves of the backscattering from high Z materials. Methods: Radiochromic EBT3 films (Ashland ISP, Wayne, USA) were wrapped around a PET hollow cylinder with a diameter of 41.5 mm and fixed upon the surface of a lead block. The setup was immersed in water and exposed to a dose of 2 Gy at 6MV acceleration voltage using a Siemens Primus linear accelerator. Water reference measurements were undertaken under equal conditions. An Epson Expression 10000 XL flatbed scanner (Epson, Suwa, Japan) with a preset resolution of 72 dpi was used for digitization. Results: The dose enhancement could be measured with a high resolution of measurement points along the axis normal to the lead surface. A dose enhancement of 70 % was measured at a distance of 134 μm from the lead surface. The data has been compared with results presented by Das et al (Med. Phys. 16(3) (1989)) and is consistent within the uncertainty of the measurements. The results are in consistence with the results from time-tested EBT3 setups, i.e. normal-to-beam EBT3 film stacks and parallel to beam EBT3-films. Conclusion: The cylindrical film setup offers a powerful tool for surface measurements. The advantages of a stacked film setup and a parallel to beam setup could be combined.

  9. Study of the penumbra for high-energy photon beams with Gafchromic™ EBT2 films

    NASA Astrophysics Data System (ADS)

    Oh, Se An; Kang, Min Kyu; Yea, Ji Woon; Kim, Sung Kyu; Oh, Young Kee

    2012-06-01

    The penumbra has a major impact on obtaining uniformity of isodose distributions in radiation therapy. The penumbra phenomena of intensity-modulated radiation therapy (IMRT) techniques using multi-leaf collimators (MLCs) has an impact on the dose distributions in the border of the target volumes and the MLC. The aim of this study was to determine the impact of high photon energy (6 MV, 10 MV) on the penumbra for various depths and field sizes by using the Pencil Beam Convolution algorithms (eclipse 8.6) and self-developing Gafchromic™ EBT2 film. For dose calculations and EBT2 measurements, we used an acryl phantom with dimensions 20 × 20 × 20 cm3. The 200 cGy dose was delivered to the central depth (10 cm) of the acryl phantom. The result of this study was that increased energy, field size and depth are rise to an increased penumbra (20% ˜ 80%) width. For a 6 MV photon energy, the penumbra widths (20%-80%) at 1.5 cm, 5 cm, and 10 cm depths were 4.2 mm, 4.4 mm, and 5.7 mm for the eclipse calculations and 2.9 mm, 4.1 mm, and 4.2 mm for the EBT2 film measurements for 10 × 10 cm2 field sizes, respectively. For a 10 MV photon energy, the penumbra widths were 4.5 mm, 4.7 mm, and 6.2 mm for eclipse calculations and 4.1 mm, 4.6 mm, and 4.9 mm for EBT2 film measurements, respectively. As the field size was changed to 3 cm, 5 cm, 7 cm, 10 cm, and 15 cm, the penumbra widths changed to 5.1 mm, 5.3 mm, 5.6 mm, 5.9 mm, and 6.1 mm for eclipse calculations and 2.9 mm, 3.3 mm, 3.6 mm, 4.2 mm, and 5.1 mm for EBT2 measurements, respectively, for 10 cm depths for 6 MV photon energies. In this study, compared to the 10 MV photon energy, the 6 MV photon energy was preferred in treatments such as the 3D conformal radiation therapy and the IMRT for critical organs near the target volume.

  10. Calibration of EBT2 film using a red-channel PDD method in combination with a modified three-channel technique

    SciTech Connect

    Chang, Liyun; Ho, Sheng-Yow; Lee, Tsair-Fwu; Yeh, Shyh-An; Ding, Hueisch-Jy; Chen, Pang-Yu

    2015-10-15

    Purpose: Ashland Inc. EBT2 and EBT3 films are widely used in quality assurance for radiation therapy; however, there remains a relatively high degree of uncertainty [B. Hartmann, M. Martisikova, and O. Jakel, “Homogeneity of Gafchromic EBT2 film,” Med. Phys. 37, 1753–1756 (2010)]. Micke et al. (2011) recently improved the spatial homogeneity using all color channels of a flatbed scanner; however, van Hoof et al. (2012) pointed out that the corrected nonuniformity still requires further investigation for larger fields. To reduce the calibration errors and the uncertainty, the authors propose a new red-channel percentage-depth-dose method in combination with a modified three-channel technique. Methods: For the ease of comparison, the EBT2 film image used in the authors’ previous study (2012) was reanalyzed using different approaches. Photon beams of 6-MV were delivered to two different films at two different beam on times, resulting in the absorption doses of ranging from approximately 30 to 300 cGy at the vertical midline of the film, which was set to be coincident with the central axis of the beam. The film was tightly sandwiched in a 30{sup 3}-cm{sup 3} polystyrene phantom, and the pixel values for red, green, and blue channels were extracted from 234 points on the central axis of the beam and compared with the corresponding depth doses. The film was first calibrated using the multichannel method proposed by Micke et al. (2010), accounting for nonuniformities in the scanner. After eliminating the scanner and dose-independent nonuniformities, the film was recalibrated via the dose-dependent optical density of the red channel and fitted to a power function. This calibration was verified via comparisons of the dose profiles extracted from the films, where three were exposed to a 60° physical wedge field and three were exposed to composite fields, and all of which were measured in a water phantom. A correction for optical attenuation was implemented, and

  11. Extraction of electron beam dose parameters from EBT2 film data scored in a mini phantom.

    PubMed

    O'Reilly, Dedri; Smit, Cobus J L; du Plessis, Freek C P

    2013-09-01

    Quality assurance of medical linear accelerators includes dosimetric parameter measurement of therapeutic electron beams e.g. relative dose at a depth of 80% (R₈₀). This parameter must be within a tolerance of 0.2 cm of the declared value. Cumbersome water tank measurements can be regarded as a benchmark to measure electron depth dose curves. A mini-phantom was designed and built, in which a strip of GAFCHROMIC® EBT2 film could be encased tightly for electron beam depth dose measurement. Depth dose data were measured for an ELEKTA Sl25 MLC, ELEKTA Precise, and ELEKTA Synergy (Elekta Oncology Systems, Crawley, UK) machines. The electron beam energy range was between 4 and 22 MeV among the machines. A 10 × 10 cm² electron applicator with 95 cm source-surface-distance was used on all the machines. 24 h after irradiation, the EBT2 film strips were scanned on Canon CanoScan N670U scanner. Afterwards, the data were analysed with in-house developed software that entailed optical density to dose conversion, and optimal fitting of the PDD data to de-noise the raw data. From the PDD data R₈₀ values were solved for and compared with acceptance values. A series of tests were also carried out to validate the use of the scanner for film Dosimetry. These tests are presented in this study. It was found that this method of R₈₀ evaluation was reliable with good agreement with benchmark water tank measurements using a commercial parallel plate ionization chamber as the radiation detector. The EBT2 film data yielded R₈₀ values that were on average 0.06 cm different from benchmark water tank measured R₈₀ values. PMID:23794059

  12. Characterization of GafChromic XR-RV2 film and comparator strip using a flatbed scanner in reflection mode

    SciTech Connect

    Mendoza-Moctezuma, A. I.

    2010-12-07

    Interventional cardiology procedures are an effective alternative for the reestablishment of correct sanguineous circulation in the heart. However, this kind of procedures exposes to the patients to a relatively high radiation doses. Usually, the surface peak skin dose is evaluated using a visual scale with a comparator strip, nevertheless, even if the comparator strip provides a simple and quick method for estimating the dose it has an uncertainty of {+-}25%. For this reason, a better evaluation method is needed. The objective of our project is to determine the surface peak skin dose of interventional cardiology procedures using GafChromic XR-RV2 film together with a commercial flatbed scanner in reflection mode. Here we report a protocol to handle GafChromic XR-RV2 film using a commercial flat bed scanner in reflection mode aiming at an uncertainty of {+-}3%.

  13. Characterization of GafChromic XR-RV2 film and comparator strip using a flatbed scanner in reflection mode

    NASA Astrophysics Data System (ADS)

    Mendoza-Moctezuma, A. I.; Aguilar, J. García; García-Garduño, O. A.

    2010-12-01

    Interventional cardiology procedures are an effective alternative for the reestablishment of correct sanguineous circulation in the heart. However, this kind of procedures exposes to the patients to a relatively high radiation doses. Usually, the surface peak skin dose is evaluated using a visual scale with a comparator strip, nevertheless, even if the comparator strip provides a simple and quick method for estimating the dose it has an uncertainty of ±25%. For this reason, a better evaluation method is needed. The objective of our project is to determine the surface peak skin dose of interventional cardiology procedures using GafChromic XR-RV2 film together with a commercial flatbed scanner in reflection mode. Here we report a protocol to handle GafChromic XR-RV2 film using a commercial flat bed scanner in reflection mode aiming at an uncertainty of ±3%.

  14. Effect of correction methods of radiochromic EBT2 films on the accuracy of IMRT QA.

    PubMed

    García-Garduño, O A; Lárraga-Gutiérrez, J M; Rodríguez-Villafuerte, M; Martínez-Dávalos, A; Rivera-Montalvo, T

    2016-01-01

    Radiochromic films are dosimeters with more favorable characteristics than other two-dimensional (2D) radiation detectors. Transmission film scanners using a linear charge-coupled device (CCD) array have a drawback of variation in response along the detector array, which may result in a nonuniform transmission of signal over the scanned image. This study uses 2D gamma index analysis to compare two methods for correcting the nonuniform response of EBT2: the first method was based on the new red:blue method proposed by the manufacturer (to compensate for small nonuniformities in the film coating) and the second method, proposed by Menegotti et al. (2008), was based on dose-dependent matrix correction factors. The gamma index analysis shows that both the methods are comparably accurate for all the criteria values used for evaluation (1 mm/1%, 2 mm/2%, 3 mm/3%). Centers around the world use both the methods to correct EBT2 local heterogeneities, but it is important to note that the former method has several advantages such as less time consumption and easy implementation. PMID:26492322

  15. A method for multichannel dosimetry with EBT3 radiochromic films

    SciTech Connect

    Pérez Azorín, Jose Fernando

    2014-06-15

    Purpose: An improved method for multichannel dosimetry is presented. This method explicitly takes into account the information provided by the unexposed image of the film. Methods: The method calculates the dose by applying a couple of perturbations to the scanned dose, one dependent and the other independent on the color channel. The method has been compared with previous multichannel and two single channel methods (red and green) against measurements using two different tests: first, five percentage depth dose profiles covering a wide range of doses; second, the dose map perpendicular to the beam axis for a 15 × 15cm{sup 2} square field. Finally, the results of 30 IMRT quality assurances tests are presented. All tests have been evaluated using the gamma analysis. Results: The coefficient of variation was found to be similar for all methods in a wide range of doses. The results of the proposed method are more in agreement with the experimental measurements and with the treatment planning system. Furthermore, the differences in the mean gamma pass rates are statistically significant. Conclusions: The improved multichannel dosimetric method is able to remove many of the common disturbances usually present in radiochromic films and improves the gamma analysis results compared with the other three methods.

  16. Characterization of a Gafchromic film for the two-dimensional profile measurement of low-energy heavy-ion beams

    NASA Astrophysics Data System (ADS)

    Yuri, Yosuke; Narumi, Kazumasa; Yuyama, Takahiro

    2016-08-01

    The feasibility of the transverse intensity distribution measurement of low-energy (keV/u range) heavy-ion beams using radiochromic films is experimentally explored. We employ a Gafchromic radiochromic film, HD-V2, whose active layer is not laminated by a surface-protection layer. The coloration response of films irradiated with several ion beams is characterized in terms of optical density (OD) by reading the films with a general-purpose scanner. To explore the energy dependence of the film response widely, the kinetic energy of the beams is varied from 1.5 keV/u to 27 MeV/u. We have found that the coloration of HD-V2 films is induced by irradiation with low-energy ion beams of the order of 10 keV/u. The range of the beams is considerably shorter than the thickness of the film's active layer. The dependence of OD response on ion species is also discussed. We demonstrate that the Gafchromic film used here is useful for measuring the intensity distribution of such low-energy ion beams.

  17. Gafchromic XR-QA2 film as a complementary dosimeter for hand-monitoring in CTF-guided biopsies.

    PubMed

    Sarmento, Sandra; Pereira, Joana; Sousa, Maria José; Cunha, Luís; Dias, Anabela G; Pereira, Miguel F; Oliveira, Augusto D; Cardoso, João V; Santos, Luis M; Gouvêa, Margarida; Lencart, Joana; Alves, João G; Santos, João A M

    2016-01-01

    Computed tomography fluoroscopy (CTF) is a useful imaging technique to guide biopsies, particularly lung biopsies, but it also has the potential for very high hand exposures, despite use of quick-check method and needle holders whenever feasible. Therefore, reliable monitoring is crucial to ensure the safe use of CTF. This is a challenge, because ring dosimeters monitor exposure only at the base of one finger, while the fingertips may be exposed to the highly collimated CT beam. In this work we have explored the possibility of using Gafchromic XR-QA2 self-developing film as a complementary dosimeter to quantify hand exposure during CTF-guided biopsies. A glove used in a previous study and designed to contain 11 TLDs was adapted to include Gafchromic strips 7 mm wide, covering the fingers. A total of 22 biopsies were successfully performed wearing this GafTLD glove under sterile gloves, and the IR reported no difficulty or reduction of dexterity while wearing it. Comparison of dose distributions obtained from digitization of the Gafchromic film strips and absolute Hp(0.07) readings from TLDs showed good agreement, despite some positional uncertainty due to relative movement. Per procedure, doses at the base of the ring finger can be as low as 3%-8% of hand dose maximum. Accumulated dose at the base of the ring finger was four times lower than the dose maximum. PMID:26894341

  18. Photon beam dosimetry in the superficial buildup region using radiochromic EBT film stack.

    PubMed

    Chiu-Tsao, Sou-Tung; Chan, Maria F

    2009-06-01

    It has been a challenge to perform accurate 2D or 3D dosimetry in the surface region with steep dose gradient for megavoltage photon beams. We developed a dosimetry method in the superficial buildup region for the 6 and 15 MV photon beams using a radiochromic EBT film stack. Eight radiochromic EBT film strips (3 x 20 x 0.024 cm3) stacked together formed a 3D dosimeter. The film stack was positioned above a polystyrene phantom and surrounded by Solid Water slabs (0.2 cm) with the top film layer at 100 cm SSD. A 10 x 10 cm2 open field was used to irradiate the film stack with 1000 MU. All films were scanned using Epson 4870 flatbed scanner with transmission mode, 48 bit color, and 150 dpi (0.017 cm pixel resolution). The pixel values were converted to doses using an established calibration curve. This method allowed dose measurement for depths from 0.012 to 0.18 cm with fine spatial resolution (0.017 cm horizontally and 0.024 cm vertically). For each energy modality, we obtained both the central axis percent depth doses and the beam profiles along the central line covering the primary field and peripheral region at each depth. The primary field doses varied steeply with depth, while those in the peripheral region were weakly dependent on depth. For the 6 MV and 15 MV photon beams, (1) the central axis percent depth doses in the eight film layers ranged from 22% to 66% and from 15% to 44%, respectively; (2) the extrapolated percent depth doses at d = 0 were 15% and 14%, respectively. Agreement with the previously reported central axis percent depth doses in this region using parallel plate thin window ion chamber and ultrathin TLD was observed. The percent depth doses and beam profiles data can be incorporated in the treatment planning system for more accurate assessment of the doses to skin and shallow tumors to accomplish more accurate calculation results in the clinical usage. PMID:19610296

  19. Improving the energy response of external beam therapy (EBT) GafChromic{sup TM} dosimetry films at low energies (≤100 keV)

    SciTech Connect

    Bekerat, H. Devic, S.; DeBlois, F.; Singh, K.; Sarfehnia, A.; Seuntjens, J.; Shih, Shelley; Yu, Xiang; Lewis, D.

    2014-02-15

    Purpose: Purpose of this work is to investigate the effects of varying the active layer composition of external beam therapy (EBT) GafChromic{sup TM} films on the energy dependence of the film, as well as try to develop a new prototype with more uniform energy response at low photon energies (⩽100 keV). Methods: First, the overall energy response (S{sub AD,} {sub W}(Q)) of different commercial EBT type film models that represent the three different generations produced to date, i.e., EBT, EBT2, and EBT3, was investigated. Pieces of each film model were irradiated to a fixed dose of 2 Gy to water for a wide range of beam qualities and the corresponding S{sub AD,} {sub W}(Q) was measured using a flatbed document scanner. Furthermore, the DOSRZnrc Monte Carlo code was used to determine the absorbed dose to water energy dependence of the film, f(Q). Moreover, the intrinsic energy dependence, k{sub bq}(Q), for each film model was evaluated using the corresponding S{sub AD,} {sub W}(Q) and f(Q). In the second part of this study, the authors investigated the effects of changing the chemical composition of the active layer on S{sub AD,} {sub W}(Q). Finally, based on these results, the film manufacturer fabricated several film prototypes and the authors evaluated their S{sub AD,} {sub W}(Q). Results: The commercial EBT film model shows an under response at all energies below 100 keV reaching 39% ± 4% at about 20 keV. The commercial EBT2 and EBT3 film models show an under response of about 27% ± 4% at 20 keV and an over response of about 16% ± 4% at 40 keV.S{sub AD,} {sub W}(Q) of the three commercial film models at low energies show strong correlation with the corresponding f{sup −1}(Q) curves. The commercial EBT3 model with 4% Cl in the active layer shows under response of 22% ± 4% at 20 keV and 6% ± 4% at about 40 keV. However, increasing the mass percent of chlorine makes the film more hygroscopic which may affect the stability of the film's readout. The

  20. Characterization of the proton beam from an IBA Cyclone 18/9 with radiochromic film EBT2

    SciTech Connect

    Sansaloni, F.; Lagares, J. I.; Arce, P.; Llop, J.; Perez, J. M.

    2012-12-19

    The use of radiochromic films is widespread in different areas of medical physics like radiotherapy and hadrontherapy; however, radiochromic films have been scarcely used in the characterization of proton or deuteron beams generated in biomedical cyclotrons. In this paper the radiochromic film EBT2 was used to study the beam size and the proton beam energy of an IBA Cyclone 18/9 cyclotron. The results indicate that the beam size can be easily measured at a very low expense; however, an accurate determination of the beam energy might require the implementation of certain experimental improvements.

  1. SU-E-T-32: An Application of GafChromic RTQA2 Film to the Patient Specified Quality Assurance

    SciTech Connect

    Peng, J; Hu, W

    2014-06-01

    Purpose: GafChromic RTQA2 film is known as a quality assurance (QA) tool for light field and radiation field verification. This study is attempted to apply the RTQA2 film to the patient specified quality assurance. Methods: Pre-irradiated and post-irradiated RTQA2 films were scanned in a reflection mode using a flatbed scanner. A plan-based dose calibration method utilized the mapping information of calculated dose image and measured film image to create a dose vs. pixel value calibration model. This model was used to calibrate the measured film image from the pixel value (gray value) image to the dose image. The dose agreement between calculated and measured dose images were analyzed using the gamma analysis. To evaluate the feasibility of this method, three clinical approved RapidArc cases (one abdomen cancer and two head-and-neck cancer patients) were tested. The tolerance of 3% dose difference and 3 mm distance to agreement (DTA) and gamma index ≤ 1 were set for the analysis. Results: The calibrated film dose image from measurement was successfully compared to the predicted dose image from the commercial treatment planning. The gamma analysis results showed good consistency. Gamma passing rates were 99.02%, 94.84%, and 98.33% for the three patients, respectively. Conclusion: The plan based calibration method has the feasibility for dose verification without shortages of film batch and development time variation.

  2. Measurement of relative output factors for the 8 and 4 mm collimators of Leksell Gamma Knife Perfexion by film dosimetry

    SciTech Connect

    Novotny, Josef Jr.; Bhatnagar, Jagdish P.; Quader, Mubina A.; Bednarz, Greg; Lunsford, L. Dade; Huq, M. Saiful

    2009-05-15

    Three types of films, Kodak EDR2, Gafchromic EBT, and Gafchromic MD-V2-55, were used to measure relative output factors of 4 and 8 mm collimators of the Leksell Gamma Knife Perfexion. The optical density to dose calibration curve for each of the film types was obtained by exposing the films to a range of known doses. Ten data points were acquired for each of the calibration curves in the dose ranges from 0 to 4 Gy, 0 to 8 Gy, and 0 to 80 Gy for Kodak EDR2, Gafchromic EBT, and Gafchromic MD-V2-55 films, respectively. For the measurement of relative output factors, five films of each film type were exposed to a known dose. All films were scanned using EPSON EXPRESSION 10000 XL scanner with 200 dpi resolution in 16 bit gray scale for EDR2 film and 48 bit color scale for Gafchromic films. The scanned images were imported in the red channel for both Gafchromic films. The background corrections from an unexposed film were applied to all films. The output factors obtained from film measurements were in a close agreement both with the Monte Carlo calculated values of 0.924 and 0.805 for 8 and 4 mm collimators, respectively. These values are provided by the vendor and used as default values in the vendor's treatment planning system. The largest differences were noted for the Kodak EDR 2 films (-2.1% and -4.5% for 8 and 4 mm collimators, respectively). The best agreement observed was for EBT Gafchromic film (-0.8% and +0.6% differences for 8 and 4 mm collimators, respectively). Based on the present values, no changes in the default relative output factor values were made in the treatment planning system.

  3. MAGAT gel and EBT2 film-based dosimetry for evaluating source plugging-based treatment plan in Gamma Knife stereotactic radiosurgery.

    PubMed

    Natanasabapathi, Gopishankar; Subbiah, Vivekanandhan; Kale, Shashank Sharad; Rath, Goura Kishor; Senthilkumaran, S; Thulkar, Sanjay; Subramani, Vellaiyan; Laviraj, M A; Bisht, Raj Kishor; Mahapatra, A K

    2012-01-01

    This work illustrates a procedure to assess the overall accuracy associated with Gamma Knife treatment planning using plugging. The main role of source plugging or blocking is to create dose falloff in the junction between a target and a critical structure. We report the use of MAGAT gel dosimeter for verification of an experimental treatment plan based on plugging. The polymer gel contained in a head-sized glass container simulated all major aspects of the treatment process of Gamma Knife radiosurgery. The 3D dose distribution recorded in the gel dosimeter was read using a 1.5T MRI scanner. Scanning protocol was: CPMG pulse sequence with 8 equidistant echoes, TR = 7 s, echo step = 14 ms, pixel size = 0.5mm × 0.5mm, and slice thickness of 2 mm. Using a calibration relationship between absorbed dose and spin-spin relaxation rate (R2), we converted R2 images to dose images. Volumetric dose comparison between treatment planning system (TPS) and gel measurement was accomplished using an in-house MATLAB-based program. The isodose overlay of the measured and computed dose distribution on axial planes was in close agreement. Gamma index analysis of 3D data showed more than 94% voxel pass rate for different tolerance criteria of 3%/2 mm, 3%/1 mm and 2%/2 mm. Film dosimetry with GAFCHROMIC EBT 2 film was also performed to compare the results with the calculated TPS dose. Gamma index analysis of film measurement for the same tolerance criteria used for gel measurement evaluation showed more than 95% voxel pass rate. Verification of gamma plan calculated dose on account of shield is not part of acceptance testing of Leksell Gamma Knife (LGK). Through this study we accomplished a volumetric comparison of dose distributions measured with a polymer gel dosimeter and Leksell GammaPlan (LGP) calculations for plans using plugging. We propose gel dosimeter as a quality assurance (QA) tool for verification of plug-based planning. PMID:23149780

  4. Dose and energy dependence of response of Gafchromic® XR-QA film for kilovoltage x-ray beams

    NASA Astrophysics Data System (ADS)

    Rampado, O.; Garelli, E.; Deagostini, S.; Ropolo, R.

    2006-06-01

    There is a growing interest in Gafchromic® films for patient dosimetry in radiotherapy and in radiology. A new model (XR-QA) with high sensitivity to low dose was tested in this study. The response of the film to different x-ray beam energies (range 28-145 kVp with various filtrations, dose range 0-100 mGy) and to visible light was investigated, together with the after exposure darkening properties. Exposed films were digitized with a commercially available, optical flatbed scanner. A single functional form for dose versus net pixel value variation has been determined for all the obtained calibration curves, with a unique fit parameter different for each of the used x-ray beams. The film response was dependent on beam energy, with higher colour variations for the beams in the range 80-140 kVp. Different sources of uncertainties in dose measurements, governed by the digitalization process, the film response uniformity and the calibration curve fit procedure, have been considered. The overall one-sigma dose measurement uncertainty depended on the beam energy and decreased with increasing absorbed dose. For doses above 10 mGy and beam energies in the range 80-140 kVp the total uncertainty was less than 5%, whereas for the 28 kVp beam the total uncertainty at 10 mGy was about 10%. The post-exposure colour variation was not negligible in the first 24 h after the exposure, with a consequent increase in the calculated dose of about 10%. Results of the analysis of the sensitivity to visible light indicated that a short exposure of this film to ambient and scanner light during the measurements will not have a significant impact on the radiation dosimetry.

  5. Dose and energy dependence of response of Gafchromic XR-QA film for kilovoltage x-ray beams.

    PubMed

    Rampado, O; Garelli, E; Deagostini, S; Ropolo, R

    2006-06-01

    There is a growing interest in Gafchromic films for patient dosimetry in radiotherapy and in radiology. A new model (XR-QA) with high sensitivity to low dose was tested in this study. The response of the film to different x-ray beam energies (range 28-145 kVp with various filtrations, dose range 0-100 mGy) and to visible light was investigated, together with the after exposure darkening properties. Exposed films were digitized with a commercially available, optical flatbed scanner. A single functional form for dose versus net pixel value variation has been determined for all the obtained calibration curves, with a unique fit parameter different for each of the used x-ray beams. The film response was dependent on beam energy, with higher colour variations for the beams in the range 80-140 kVp. Different sources of uncertainties in dose measurements, governed by the digitalization process, the film response uniformity and the calibration curve fit procedure, have been considered. The overall one-sigma dose measurement uncertainty depended on the beam energy and decreased with increasing absorbed dose. For doses above 10 mGy and beam energies in the range 80-140 kVp the total uncertainty was less than 5%, whereas for the 28 kVp beam the total uncertainty at 10 mGy was about 10%. The post-exposure colour variation was not negligible in the first 24 h after the exposure, with a consequent increase in the calculated dose of about 10%. Results of the analysis of the sensitivity to visible light indicated that a short exposure of this film to ambient and scanner light during the measurements will not have a significant impact on the radiation dosimetry. PMID:16723772

  6. Development of the optimal radiochromic film dosimetry system for measurement of IMRT radiation beams

    NASA Astrophysics Data System (ADS)

    Baker, Jameson Todd

    The complex dose patterns that result in Intensity Modulated Radiation Therapy make the typical QA of a second calculation insufficient for ensuring safe treatment of patients. Many facilities choose to deliver the treatment to film inserted in a phantom and calculate the dose delivered as an additional check of the treatment plan. Radiochromic films allow for measurements without the use of a processor in the current digital age. International Specialty Products developed Gafchromic EBT film, which is a radiochromic film having a useful range of 1 -- 800 cGy. EBT film properties are fully analyzed including studies of uniformity, spectral absorption, exposure sensitivity, energy dependence and post exposure density growth. Dosimetric performance on commercially available digitizers is studied with specific attention on the shortcomings. Finally, a custom designed scanner is built specifically for EBT film and its unique properties. Performance of the EBT digitizer is analyzed and compared against currently available scanners.

  7. Accurate skin dose measurements using radiochromic film in clinical applications

    SciTech Connect

    Devic, S.; Seuntjens, J.; Abdel-Rahman, W.; Evans, M.; Olivares, M.; Podgorsak, E.B.; Vuong, Te; Soares, Christopher G.

    2006-04-15

    Megavoltage x-ray beams exhibit the well-known phenomena of dose buildup within the first few millimeters of the incident phantom surface, or the skin. Results of the surface dose measurements, however, depend vastly on the measurement technique employed. Our goal in this study was to determine a correction procedure in order to obtain an accurate skin dose estimate at the clinically relevant depth based on radiochromic film measurements. To illustrate this correction, we have used as a reference point a depth of 70 {mu}. We used the new GAFCHROMIC[reg] dosimetry films (HS, XR-T, and EBT) that have effective points of measurement at depths slightly larger than 70 {mu}. In addition to films, we also used an Attix parallel-plate chamber and a home-built extrapolation chamber to cover tissue-equivalent depths in the range from 4 {mu} to 1 mm of water-equivalent depth. Our measurements suggest that within the first millimeter of the skin region, the PDD for a 6 MV photon beam and field size of 10x10 cm{sup 2} increases from 14% to 43%. For the three GAFCHROMIC[reg] dosimetry film models, the 6 MV beam entrance skin dose measurement corrections due to their effective point of measurement are as follows: 15% for the EBT, 15% for the HS, and 16% for the XR-T model GAFCHROMIC[reg] films. The correction factors for the exit skin dose due to the build-down region are negligible. There is a small field size dependence for the entrance skin dose correction factor when using the EBT GAFCHROMIC[reg] film model. Finally, a procedure that uses EBT model GAFCHROMIC[reg] film for an accurate measurement of the skin dose in a parallel-opposed pair 6 MV photon beam arrangement is described.

  8. Using GafChromic film to estimate the effective dose from dental cone beam CT and panoramic radiography

    PubMed Central

    Al-Okshi, A; Nilsson, M; Petersson, A; Wiese, M; Lindh, C

    2013-01-01

    Objectives: To demonstrate the feasibility of GafChromic® XR-QA2 (ISP Corp., Wayne, NJ) as a dosemeter when performing measurements of the effective dose from three cone beam CT (CBCT) units and to compare the doses from examinations of three common dental clinical situations. A second aim was to compare the radiation doses for three digital panoramic units with the doses for the CBCT units. Methods: The CBCT units used were Veraviewepocs 3De® (J Morita MFG Corp., Kyoto, Japan), ProMax® 3D (Planmeca, Helsinki, Finland) and NewTom VGi® (Quantitative Radiology, Verona, Italy). GafChromic XR-QA2 films were placed between the selected layers of the head and neck of a tissue-equivalent human skull (RANDO® phantom; The Phantom Laboratory, Salem, NY). The exposure parameters were set using the automatic exposure control function of the units. Depending on the availability, medium and smaller field of view (FOV) scanning modes were used. The effective dose was estimated using the 2007 International Commission on Radiological Protection formalism. Results: The lowest effective dose of a CBCT unit was observed for ProMax 3D, FOV 4 × 5 cm (10 μSv), the highest for NewTom VGi, FOV 8 × 8 cm—high resolution (129 μSv). The range of effective doses for digital panoramic machines measured was 8–14 μSv. Conclusions: This study demonstrates the feasibility of using radiochromic films for dental CBCT and panoramic dosimetry. PMID:23610090

  9. Multidimensional dosimetry of {sup 106}Ru eye plaques using EBT3 films and its impact on treatment planning

    SciTech Connect

    Heilemann, G. Kostiukhina, N.

    2015-10-15

    Purpose: The purpose of this study was to establish a method to perform multidimensional radiochromic film measurements of {sup 106}Ru plaques and to benchmark the resulting dose distributions against Monte Carlo simulations (MC), microdiamond, and diode measurements. Methods: Absolute dose rates and relative dose distributions in multiple planes were determined for three different plaque models (CCB, CCA, and COB), and three different plaques per model, using EBT3 films in an in-house developed polystyrene phantom and the MCNP6 MC code. Dose difference maps were generated to analyze interplaque variations for a specific type, and for comparing measurements against MC simulations. Furthermore, dose distributions were validated against values specified by the manufacturer (BEBIG) and microdiamond and diode measurements in a water scanning phantom. Radial profiles were assessed and used to estimate dosimetric margins for a given combination of representative tumor geometry and plaque size. Results: Absolute dose rates at a reference depth of 2 mm on the central axis of the plaque show an agreement better than 5% (10%) when comparing film measurements (MCNP6) to the manufacturer’s data. The reproducibility of depth-dose profile measurements was <7% (2 SD) for all investigated detectors and plaque types. Dose difference maps revealed minor interplaque deviations for a specific plaque type due to inhomogeneities of the active layer. The evaluation of dosimetric margins showed that for a majority of the investigated cases, the tumor was not completely covered by the 100% isodose prescribed to the tumor apex if the difference between geometrical plaque size and tumor base ≤4 mm. Conclusions: EBT3 film dosimetry in an in-house developed phantom was successfully used to characterize the dosimetric properties of different {sup 106}Ru plaque models. The film measurements were validated against MC calculations and other experimental methods and showed a good agreement with

  10. Measurement of absorbed dose to water around an electronic brachytherapy source. Comparison of two dosimetry systems: lithium formate EPR dosimeters and radiochromic EBT2 film

    NASA Astrophysics Data System (ADS)

    Adolfsson, Emelie; White, Shane; Landry, Guillaume; Lund, Eva; Gustafsson, Håkan; Verhaegen, Frank; Reniers, Brigitte; Carlsson Tedgren, Åsa; Alm Carlsson, Gudrun

    2015-05-01

    Interest in high dose rate (HDR) electronic brachytherapy operating at 50 kV is increasing. For quality assurance it is important to identify dosimetry systems that can measure the absorbed doses in absolute terms which is difficult in this energy region. In this work a comparison is made between two dosimetry systems, EPR lithium formate dosimeters and radiochromic EBT2 film. Both types of dosimeters were irradiated simultaneously in a PMMA phantom using the Axxent EBS. Absorbed dose to water was determined at distances of 10 mm, 30 mm and 50 mm from the EBS. Results were traceable to different primary standards as regards to absorbed dose to water (EPR) and air kerma (EBT2). Monte Carlo simulations were used in absolute terms as a third estimate of absorbed dose to water. Agreement within the estimated expanded (k = 2) uncertainties (5% (EPR), 7% (EBT2)) was found between the results at 30 mm and 50 mm from the x-ray source. The same result was obtained in 4 repetitions of irradiation, indicating high precision in the measurements with both systems. At all distances, agreement between EPR and Monte Carlo simulations was shown as was also the case for the film measurements at 30mm and 50mm. At 10mm the geometry for the film measurements caused too large uncertainty in measured values depending on the exact position (within sub-mm distances) of the EBS and the 10 mm film results were exculded from comparison. This work has demonstrated good performance of the lithium formate EPR dosimetry system in accordance with earlier experiments at higher photon energies (192Ir HDR brachytherapy). It was also highlighted that there might be issues regarding the energy dependence and intrinsic efficiency of the EBT2 film that need to be considered for measurements using low energy sources.

  11. Poster — Thur Eve — 39: Feasibility of Commissioning HybridArc with the Delta 4 two plane diode phantom: comparisons with Gafchromic Film

    SciTech Connect

    Bojechko, C.; Ploquin, N.; Hudson, A.; Sayous, Y.

    2014-08-15

    HybridArc is a relatively novel radiation therapy technique which combines optimized dynamic conformai arcs (DCA) and intensity modulated radiation therapy (IMRT). HybridArc has possible dosimetry and efficiency advantages over stand alone DCA and IMRT treatments and can be readily implemented on any linac capable of DCA and IMRT, giving strong motivation to commission the modality. The Delta4 phantom (Scandidos, Uppsala, Sweden) has been used for IMRT and VMAT clinical dosimetric verification making it a candidate for HybridArc commissioning. However the HybridArc modality makes use of several non co-planar arcs which creates setup issues due to the geometry of the Delta4, resulting in possible phantom gantry collisions for plans with non-zero couch angles. An analysis was done determining the feasibility of using the Delta4 fixed at 0° couch angle compared with results obtained using Gafchromic ETB2 film (Ashland, Covington Kentucky) in an anthropomorphic phantom at the planned couch angles. A gamma index analysis of the measured and planned dose distributions was done using Delta4 and DoseLab Pro (Mobius Medical Systems, Houston Texas) software. For both arc and IMRT sub-fields there is reasonable correlation between the gamma index found from the Delta4 and Gafchromic film. All results show the feasibility of using the Delta4 for HybridArc commissioning.

  12. How flatbed scanners upset accurate film dosimetry.

    PubMed

    van Battum, L J; Huizenga, H; Verdaasdonk, R M; Heukelom, S

    2016-01-21

    Film is an excellent dosimeter for verification of dose distributions due to its high spatial resolution. Irradiated film can be digitized with low-cost, transmission, flatbed scanners. However, a disadvantage is their lateral scan effect (LSE): a scanner readout change over its lateral scan axis. Although anisotropic light scattering was presented as the origin of the LSE, this paper presents an alternative cause. Hereto, LSE for two flatbed scanners (Epson 1680 Expression Pro and Epson 10000XL), and Gafchromic film (EBT, EBT2, EBT3) was investigated, focused on three effects: cross talk, optical path length and polarization. Cross talk was examined using triangular sheets of various optical densities. The optical path length effect was studied using absorptive and reflective neutral density filters with well-defined optical characteristics (OD range 0.2-2.0). Linear polarizer sheets were used to investigate light polarization on the CCD signal in absence and presence of (un)irradiated Gafchromic film. Film dose values ranged between 0.2 to 9 Gy, i.e. an optical density range between 0.25 to 1.1. Measurements were performed in the scanner's transmission mode, with red-green-blue channels. LSE was found to depend on scanner construction and film type. Its magnitude depends on dose: for 9 Gy increasing up to 14% at maximum lateral position. Cross talk was only significant in high contrast regions, up to 2% for very small fields. The optical path length effect introduced by film on the scanner causes 3% for pixels in the extreme lateral position. Light polarization due to film and the scanner's optical mirror system is the main contributor, different in magnitude for the red, green and blue channel. We concluded that any Gafchromic EBT type film scanned with a flatbed scanner will face these optical effects. Accurate dosimetry requires correction of LSE, therefore, determination of the LSE per color channel and dose delivered to the film. PMID:26689962

  13. How flatbed scanners upset accurate film dosimetry

    NASA Astrophysics Data System (ADS)

    van Battum, L. J.; Huizenga, H.; Verdaasdonk, R. M.; Heukelom, S.

    2016-01-01

    Film is an excellent dosimeter for verification of dose distributions due to its high spatial resolution. Irradiated film can be digitized with low-cost, transmission, flatbed scanners. However, a disadvantage is their lateral scan effect (LSE): a scanner readout change over its lateral scan axis. Although anisotropic light scattering was presented as the origin of the LSE, this paper presents an alternative cause. Hereto, LSE for two flatbed scanners (Epson 1680 Expression Pro and Epson 10000XL), and Gafchromic film (EBT, EBT2, EBT3) was investigated, focused on three effects: cross talk, optical path length and polarization. Cross talk was examined using triangular sheets of various optical densities. The optical path length effect was studied using absorptive and reflective neutral density filters with well-defined optical characteristics (OD range 0.2-2.0). Linear polarizer sheets were used to investigate light polarization on the CCD signal in absence and presence of (un)irradiated Gafchromic film. Film dose values ranged between 0.2 to 9 Gy, i.e. an optical density range between 0.25 to 1.1. Measurements were performed in the scanner’s transmission mode, with red-green-blue channels. LSE was found to depend on scanner construction and film type. Its magnitude depends on dose: for 9 Gy increasing up to 14% at maximum lateral position. Cross talk was only significant in high contrast regions, up to 2% for very small fields. The optical path length effect introduced by film on the scanner causes 3% for pixels in the extreme lateral position. Light polarization due to film and the scanner’s optical mirror system is the main contributor, different in magnitude for the red, green and blue channel. We concluded that any Gafchromic EBT type film scanned with a flatbed scanner will face these optical effects. Accurate dosimetry requires correction of LSE, therefore, determination of the LSE per color channel and dose delivered to the film.

  14. Evaluation of the incident directional dependence of radiochromic film by use of Monte Carlo simulation and measurement.

    PubMed

    Yuasa, Yuki; Kawamura, Shinji; Shiinoki, Takehiro; Uehara, Takuya; Koike, Masahiro; Kanzaki, Ryuji; Hanazawa, Hideki; Takahashi, Shotaro; Shibuya, Keiko

    2016-07-01

    In high-precision radiotherapy, absolute and relative doses are evaluated for patient-specific intensity-modulated radiation therapy (IMRT) quality assurance (QA). In our institution, we use GAFCHROMIC EBT3 (EBT3) for relative dose evaluation in IMRT QA. We usually use two directional film configurations, which are in the axial and sagittal planes. The QA in our institution shows some differences between the gamma pass rates in the axial and sagittal directions. The purpose of this study was to evaluate the incident directional dependence of EBT3 by using the percent depth dose (PDD) and the off-center ratio (OCR) between EBT3 films positioned perpendicular to the beam axis and along the beam axis. Furthermore, we compared the PDD in EBT3 films positioned perpendicular to the beam axis and the PDD by using an ionization chamber. In addition, PDDs in water phantoms with and without EBT3 films were calculated by Monte Carlo simulation. The results showed that the PDD in EBT3 films positioned perpendicular to the beam axis increased with the depth from the phantom surface. Monte Carlo simulation showed the same trend as did the film measurements. The OCR results were slightly different at dose levels below 20 %. The OCR in EBT3 films positioned along the beam axis was higher than that perpendicular to the beam axis. Thus, we conclude that EBT3 film has incident directional dependence. In IMRT QA, the gamma analysis results may be affected by the incident directional dependence of EBT3 film. PMID:27010193

  15. Comparative dose evaluations between XVI and OBI cone beam CT systems using Gafchromic XRQA2 film and nanoDot optical stimulated luminescence dosimeters

    SciTech Connect

    Giaddui, Tawfik; Cui Yunfeng; Galvin, James; Yu Yan; Xiao Ying

    2013-06-15

    Purpose: To investigate the effect of energy (kVp) and filters (no filter, half Bowtie, and full Bowtie) on the dose response curves of the Gafchromic XRQA2 film and nanoDot optical stimulated luminescence dosimeters (OSLDs) in CBCT dose fields. To measure surface and internal doses received during x-ray volume imager (XVI) (Version R4.5) and on board imager (OBI) (Version 1.5) CBCT imaging protocols using these two types of dosimeters. Methods: Gafchromic XRQA2 film and nanoDot OSLD dose response curves were generated at different kV imaging settings used by XVI (software version R4.5) and OBI (software version 1.5) CBCT systems. The settings for the XVI system were: 100 kVp/F0 (no filter), 120 kVp/F0, and 120 kVp/F1 (Bowtie filter), and for the OBI system were: 100 kVp/full fan, 125 kVp/full fan, and 125 kVp/half fan. XRQA2 film was calibrated in air to air kerma levels between 0 and 11 cGy and scanned using reflection scanning mode with the Epson Expression 10000 XL flat-bed document scanner. NanoDot OSLDs were calibrated on phantom to surface dose levels between 0 and 14 cGy and read using the inLight{sup TM} MicroStar reader. Both dosimeters were used to measure in field surface and internal doses in a male Alderson Rando Phantom. Results: Dose response curves of XRQA2 film and nanoDot OSLDs at different XVI and OBI CBCT settings were reported. For XVI system, the surface dose ranged between 0.02 cGy in head region during fast head and neck scan and 4.99 cGy in the chest region during symmetry scan. On the other hand, the internal dose ranged between 0.02 cGy in the head region during fast head and neck scan and 3.17 cGy in the chest region during chest M20 scan. The average (internal and external) dose ranged between 0.05 cGy in the head region during fast head and neck scan and 2.41 cGy in the chest region during chest M20 scan. For OBI system, the surface dose ranged between 0.19 cGy in head region during head scan and 4.55 cGy in the pelvis region during

  16. External Beam Therapy (EBT)

    MedlinePlus

    ... Physician Resources Professions Site Index A-Z External Beam Therapy (EBT) External beam therapy (EBT) is a ... follow-up should I expect? What is external beam therapy and how is it used? External beam ...

  17. SU-E-T-44: Angular Dependence of Surface Dose Enhancement Measured On Several Inhomogeneities Using Radiochromic EBT3 Films

    SciTech Connect

    Jansen, A; Schoenfeld, A; Poppinga, D; Chofor, N; Poppe, B

    2014-06-01

    Purpose: The quantification of the relative surface dose enhancement in dependence on the angle of incidence and the atomic number Z of the surface material. Methods: Experiments were performed with slabs made of aluminum, titanium, copper, silver, dental gold and lead. The metal slabs with equal sizes of 1.0×8.0×8.8mm{sup 3} were embedded in an Octavius 4D phantom (PTW Freiburg, Germany). Radiochromic EBT3 films were used to measure the surface dose for angles of incidence ranging from 0° to 90°. The setup with the metals slabs at the isocenter was irradiated with acceleration voltages of 6MV and 10MV. Water reference measurements were taken under equal conditions. Results: The surface dose enhancement is highest for angles of incidence below 30° and drops significantly for higher. The surface dose enhancement produced by lead and dental gold at 6MV showed a peak of 65%. At 90°, the surface dose enhancement dropped to 15% for both materials. The surface dose enhancements for silver, copper, titanium and aluminum were 45%, 32%, 22% and 12% at 0°, respectively. At an angle of incidence of 80°, the values dropped to 22%, 18%, 12% und 6%. The values for 10MV were very similar. Lead and dental gold showed peaks of 65% und 60%. Their values dropped to 18% at an angle of 90°. The surface dose enhancements for silver, copper, titanium and aluminum were 45%, 30%, 20% and 8% at 0°. At 80° the values dropped to 30%, 20%, 12% and 5%. A dependence of the magnitude of the surface dose enhancement on the atomic number of the surface material can be seen, which is in consistence with literature. Conclusion: The results show that the surface dose enhancements near implant materials with high Z-values should be taken into consideration in radio therapy, even when the angle of incidence is flat.

  18. Precise radiochromic film dosimetry using a flat-bed document scanner

    SciTech Connect

    Devic, Slobodan; Seuntjens, Jan; Sham, Edwin; Podgorsak, Ervin B.; Schmidtlein, C. Ross; Kirov, Assen S.; Soares, Christopher G.

    2005-07-15

    In this study, a measurement protocol is presented that improves the precision of dose measurements using a flat-bed document scanner in conjunction with two new GafChromic registered film models, HS and Prototype A EBT exposed to 6 MV photon beams. We established two sources of uncertainties in dose measurements, governed by measurement and calibration curve fit parameters contributions. We have quantitatively assessed the influence of different steps in the protocol on the overall dose measurement uncertainty. Applying the protocol described in this paper on the Agfa Arcus II flat-bed document scanner, the overall one-sigma dose measurement uncertainty for an uniform field amounts to 2% or less for doses above around 0.4 Gy in the case of the EBT (Prototype A), and for doses above 5 Gy in the case of the HS model GafChromic registered film using a region of interest 2x2 mm{sup 2} in size.

  19. Precise radiochromic film dosimetry using a flat-bed document scanner.

    PubMed

    Devic, Slobodan; Seuntjens, Jan; Sham, Edwin; Podgorsak, Ervin B; Schmidtlein, C Ross; Kirov, Assen S; Soares, Christopher G

    2005-07-01

    In this study, a measurement protocol is presented that improves the precision of dose measurements using a flat-bed document scanner in conjunction with two new GafChromic film models, HS and Prototype A EBT exposed to 6 MV photon beams. We established two sources of uncertainties in dose measurements, governed by measurement and calibration curve fit parameters contributions. We have quantitatively assessed the influence of different steps in the protocol on the overall dose measurement uncertainty. Applying the protocol described in this paper on the Agfa Arcus II flat-bed document scanner, the overall one-sigma dose measurement uncertainty for an uniform field amounts to 2% or less for doses above around 0.4 Gy in the case of the EBT (Prototype A), and for doses above 5 Gy in the case of the HS model GafChromic film using a region of interest 2 X 2 mm2 in size. PMID:16121579

  20. Developing new extension of GafChromic RTQA2 film to patient quality assurance field using a plan-based calibration method

    NASA Astrophysics Data System (ADS)

    Peng, Jiayuan; Zhang, Zhen; Wang, Jiazhou; Xie, Jiang; Chen, Junchao; Hu, Weigang

    2015-10-01

    GafChromic RTQA2 film is a type of radiochromic film designed for light field and radiation field alignment. The aim of this study is to extend the application of RTQA2 film to the measurement of patient specific quality assurance (QA) fields as a 2D relative dosimeter. Pre-irradiated and post-irradiated RTQA2 films were scanned in reflection mode using a flatbed scanner. A plan-based calibration (PBC) method utilized the mapping information of the calculated dose image and film grayscale image to create a dose versus pixel value calibration model. This model was used to calibrate the film grayscale image to the film relative dose image. The dose agreement between calculated and film dose images were analyzed by gamma analysis. To evaluate the feasibility of this method, eight clinically approved RapidArc cases (one abdomen cancer and seven head-and-neck cancer patients) were tested using this method. Moreover, three MLC gap errors and two MLC transmission errors were introduced to eight Rapidarc cases respectively to test the robustness of this method. The PBC method could overcome the film lot and post-exposure time variations of RTQA2 film to get a good 2D relative dose calibration result. The mean gamma passing rate of eight patients was 97.90%  ±  1.7%, which showed good dose consistency between calculated and film dose images. In the error test, the PBC method could over-calibrate the film, which means some dose error in the film would be falsely corrected to keep the dose in film consistent with the dose in the calculated dose image. This would then lead to a false negative result in the gamma analysis. In these cases, the derivative curve of the dose calibration curve would be non-monotonic which would expose the dose abnormality. By using the PBC method, we extended the application of more economical RTQA2 film to patient specific QA. The robustness of the PBC method has been improved by analyzing the monotonicity of the derivative of the

  1. Developing new extension of GafChromic RTQA2 film to patient quality assurance field using a plan-based calibration method.

    PubMed

    Peng, Jiayuan; Zhang, Zhen; Wang, Jiazhou; Xie, Jiang; Chen, Junchao; Hu, Weigang

    2015-10-01

    GafChromic RTQA2 film is a type of radiochromic film designed for light field and radiation field alignment. The aim of this study is to extend the application of RTQA2 film to the measurement of patient specific quality assurance (QA) fields as a 2D relative dosimeter.Pre-irradiated and post-irradiated RTQA2 films were scanned in reflection mode using a flatbed scanner. A plan-based calibration (PBC) method utilized the mapping information of the calculated dose image and film grayscale image to create a dose versus pixel value calibration model. This model was used to calibrate the film grayscale image to the film relative dose image. The dose agreement between calculated and film dose images were analyzed by gamma analysis. To evaluate the feasibility of this method, eight clinically approved RapidArc cases (one abdomen cancer and seven head-and-neck cancer patients) were tested using this method. Moreover, three MLC gap errors and two MLC transmission errors were introduced to eight Rapidarc cases respectively to test the robustness of this method.The PBC method could overcome the film lot and post-exposure time variations of RTQA2 film to get a good 2D relative dose calibration result. The mean gamma passing rate of eight patients was 97.90%  ±  1.7%, which showed good dose consistency between calculated and film dose images. In the error test, the PBC method could over-calibrate the film, which means some dose error in the film would be falsely corrected to keep the dose in film consistent with the dose in the calculated dose image. This would then lead to a false negative result in the gamma analysis. In these cases, the derivative curve of the dose calibration curve would be non-monotonic which would expose the dose abnormality.By using the PBC method, we extended the application of more economical RTQA2 film to patient specific QA. The robustness of the PBC method has been improved by analyzing the monotonicity of the derivative of the calibration

  2. Technical Note: Influence of the phantom material on the absorbed-dose energy dependence of the EBT3 radiochromic film for photons in the energy range 3 keV–18 MeV

    SciTech Connect

    Hermida-López, M.; Lüdemann, L.; Flühs, A.; Brualla, L.

    2014-11-01

    Purpose: Water is the reference medium for radiation therapy dosimetry, but for film dosimetry it is more practical to use a solid phantom. As the composition of solid phantoms differs from that of water, the energy dependence of film exposed within solid phantoms may also differ. The energy dependence of a radiochromic film for a given beam quality Q (energy for monoenergetic beams) has two components: the intrinsic energy dependence and the absorbed-dose energy dependence f(Q), the latter of which can be calculated through a Monte Carlo simulation of radiation transport. The authors used Monte Carlo simulations to study the influence of the phantom material on the f(Q) of the EBT3 radiochromic film (Ashland Specialty Ingredients, Wayne, NJ) for photon beams with energies between 3 keV and 18 MeV. Methods: All simulations were carried out with the general-purpose Monte Carlo code PENELOPE 2011. The geometrical model consisted of a cylindrical phantom, with the film positioned at different depths depending on the initial photon energy. The authors simulated monoenergetic parallel photon beams and x-ray beams from a superficial therapy system. To validate their choice of simulation parameters, they also calculated f(Q) for older film models, EBT and EBT2, comparing with published results. In addition to water, they calculated f(Q) of the EBT3 film for solid phantom materials commonly used for film dosimetry: RW1 and RW3 (PTW-Freiburg, Freiburg, Germany), Solid Water (Gammex-RMI, Madison, WI), and PMMA. Finally, they combined their calculated f(Q) with published overall energy response data to obtain the intrinsic energy dependence of the EBT3 film in water. Results: The calculated f(Q) for EBT and EBT2 films was statistically compatible with previously published data. Between 10 keV and 18 MeV, the variation found in f(Q) of the EBT3 film for water was within 2.3%, with a standard statistical uncertainty less than 1%. If the quantity dose-to-water in the phantom is

  3. A preliminary study on using the radiochromic film for 2D beam profile QC/QA at the THOR BNCT facility.

    PubMed

    Hsiao, Ming-Chen; Chen, Wei-Lin; Tsai, Pi-En; Huang, Chun-Kai; Liu, Yuan-Hao; Liu, Hong-Ming; Jiang, Shiang-Huei

    2011-12-01

    The GAFCHROMIC(®) EBT2 dosimetry film has been studied as a rapid QC/QA tool for 2D dose profile mapping in the BNCT beam at THOR. The pixel values of the EBT2 film image were converted to the 2D dose profile using a dose calibration curve obtained by 6-MV X-ray. The reproducibility of the 2D dose profile measured using the EBT2 film in the PMMA phantom was preliminarily found to be acceptable with uncertainties within about ±2 to ±3.5%. It is found that the EBT2 measured dose profile consisted of both gamma-ray components and neutron contributions. Therefore, the dose profile measured using the EBT2 film is significantly different from the neutron flux profile measured using the indirect neutron radiography method. Further study of the influence of neutrons to the response of the EBT2 film is indispensible for the absolute dose profile determination in a BNCT beam. PMID:21570854

  4. Poster — Thur Eve — 30: 4D VMAT dose calculation methodology to investigate the interplay effect: experimental validation using TrueBeam Developer Mode and Gafchromic film

    SciTech Connect

    Teke, T; Milette, MP; Huang, V; Thomas, SD

    2014-08-15

    The interplay effect between the tumor motion and the radiation beam modulation during a VMAT treatment delivery alters the delivered dose distribution from the planned one. This work present and validate a method to accurately calculate the dose distribution in 4D taking into account the tumor motion, the field modulation and the treatment starting phase. A QUASAR™ respiratory motion phantom was 4D scanned with motion amplitude of 3 cm and with a 3 second period. A static scan was also acquired with the lung insert and the tumor contained in it centered. A VMAT plan with a 6XFFF beam was created on the averaged CT and delivered on a Varian TrueBeam and the trajectory log file was saved. From the trajectory log file 10 VMAT plans (one for each breathing phase) and a developer mode XML file were created. For the 10 VMAT plans, the tumor motion was modeled by moving the isocentre on the static scan, the plans were re-calculated and summed in the treatment planning system. In the developer mode, the tumor motion was simulated by moving the couch dynamically during the treatment. Gafchromic films were placed in the QUASAR phantom static and irradiated using the developer mode. Different treatment starting phase were investigated (no phase shift, maximum inhalation and maximum exhalation). Calculated and measured isodose lines and profiles are in very good agreement. For each starting phase, the dose distribution exhibit significant differences but are accurately calculated with the methodology presented in this work.

  5. Evaluation of two-dimensional bolus effect of immobilization/support devices on skin doses: A radiochromic EBT film dosimetry study in phantom

    SciTech Connect

    Chiu-Tsao, Sou-Tung; Chan, Maria F.

    2010-07-15

    Purpose: In this study, the authors have quantified the two-dimensional (2D) perspective of skin dose increase using EBT film dosimetry in phantom in the presence of patient immobilization devices during conventional and IMRT treatments. Methods: For 6 MV conventional photon field, the authors evaluated and quantified the 2D bolus effect on skin doses for six different common patient immobilization/support devices, including carbon fiber grid with Mylar sheet, Orfit carbon fiber base plate, balsa wood board, Styrofoam, perforated AquaPlast sheet, and alpha-cradle. For 6 and 15 MV IMRT fields, a stack of two film layers positioned above a solid phantom was exposed at the air interface or in the presence of a patient alpha-cradle. All the films were scanned and the pixel values were converted to doses based on an established calibration curve. The authors determined the 2D skin dose distributions, isodose curves, and cross-sectional profiles at the surface layers with or without the immobilization/support device. The authors also generated and compared the dose area histograms (DAHs) and dose area products from the 2D skin dose distributions. Results: In contrast with 20% relative dose [(RD) dose relative to d{sub max} on central axis] at 0.0153 cm in the film layer for 6 MV 10x10 cm{sup 2} open field, the average RDs at the same depth in the film layer were 71%, 69%, 55%, and 57% for Orfit, balsa wood, Styrofoam, and alpha-cradle, respectively. At the same depth, the RDs were 54% under a strut and 26% between neighboring struts of a carbon fiber grid with Mylar sheet, and between 34% and 56% for stretched perforated AquaPlast sheet. In the presence of the alpha-cradle for the 6 MV (15 MV) IMRT fields, the hot spot doses at the effective measurement depths of 0.0153 and 0.0459 cm were 140% and 150% (83% and 89%), respectively, of the isocenter dose. The enhancement factor was defined as the ratio of a given DAH parameter (minimum dose received in a given area) with

  6. SU-E-J-194: Dynamic Tumor Tracking End-To-End Testing Using a 4D Thorax Phantom and EBT3 Films

    SciTech Connect

    Su, Z; Wu, J; Li, Z; Mamalui-Hunter, M

    2014-06-01

    Purpose: To quantify the Vero linac dosimetric accuracy of the tumor dynamic tracking treatment using EBT3 film embedded in a 4D thorax phantom. Methods: A dynamic thorax phantom with tissue equivalent materials and a film insert were used in this study. The thorax phantom was scanned in 4DCT mode with a viscoil embedded in its film insert composed of lung equivalent material. Dynamic tracking planning was performed using the 50% phase CT set with 5 conformal beams at gantry angles of 330, 15, 60, 105 and 150 degrees. Each field is a 3cm by 3cm square centered at viscoil since there was no solid mass target. Total 3 different 1–2cos4 motion profiles were used with varied motion magnitude and cycle frequency. Before treatment plan irradiation, a 4D motion model of the target was established using a series of acquired fluoroscopic images and infrared markers motion positions. During irradiation, fluoroscopic image monitoring viscoil motion was performed to verify model validity. The irradiated films were scanned and the dose maps were compared to the planned Monte Carlo dose distributions. Gamma analyses using 3%–3mm, 2%–3mm, 3%–2mm, 2%–2mm criteria were performed and presented. Results: For each motion pattern, a 4D motion model was built successfully and the target tracking performance was verified with fluoroscopic monitoring of the viscoil motion and its model predicted locations. The film gamma analysis showed the average pass rates among the 3 motion profiles are 98.14%, 96.2%, 91.3% and 85.61% for 3%–3mm, 2%–3mm, 3%–2mm, 2%–2mm criteria. Conclusion: Target dynamic tracking was performed using patient-like breathing patterns in a 4D thorax phantom with EBT3 film insert and a viscoil. There was excellent agreement between acquired and planned dose distributions for all three target motion patterns. This study performed end-to-end testing and verified the treatment accuracy of tumor dynamic tracking.

  7. Absorption spectra of irradiated XRCT radiochromic film

    NASA Astrophysics Data System (ADS)

    Butson, Martin J.; Cheung, Tsang; Yu, Peter K. N.

    2006-06-01

    Gafchromic XRCT radiochromic film is a self-developing high sensitivity radiochromic film product which can be used for assessment of delivered radiation doses which could match applications such as computed tomography (CT) dosimetry. The film automatically changes colour upon irradiation changing from a yellow to green/brown colour. The absorption spectra of Gafchromic XRCT radiochromic film as measured with reflectance spectrophotometry have been investigated to analyse the dosimetry characteristics of the film. Results show two main absorption peaks produced from irradiation located at 636 nm and 585 nm. This is similar to EBT Gafchromic film. A high level of sensitivity is found for this film with a 1 cGy applied dose producing an approximate net optical density change of 0.3 at 636 nm. This high sensitivity combined with its relatively energy independent nature around the 100 kVp to 150 kVp x-ray energy range provides a unique enhancement in dosimetric measurement capabilities over currently available dosimetry films for CT applications.

  8. Optimizing the dynamic range extension of a radiochromic film dosimetry system

    SciTech Connect

    Devic, Slobodan; Tomic, Nada; Soares, Christopher G.; Podgorsak, Ervin B.

    2009-02-15

    The authors present a radiochromic film dosimetry protocol for a multicolor channel radiochromic film dosimetry system consisting of the external beam therapy (EBT) model GAFCHROMIC film and the Epson Expression 1680 flat-bed document scanner. Instead of extracting only the red color channel, the authors are using all three color channels in the absorption spectrum of the EBT film to extend the dynamic dose range of the radiochromic film dosimetry system. By optimizing the dose range for each color channel, they obtained a system that has both precision and accuracy below 1.5%, and the optimized ranges are 0-4 Gy for the red channel, 4-50 Gy for the green channel, and above 50 Gy for the blue channel.

  9. I-125 ROPES eye plaque dosimetry: Validation of a commercial 3D ophthalmic brachytherapy treatment planning system and independent dose calculation software with GafChromic{sup ®} EBT3 films

    SciTech Connect

    Poder, Joel; Corde, Stéphanie

    2013-12-15

    Purpose: The purpose of this study was to measure the dose distributions for different Radiation Oncology Physics and Engineering Services, Australia (ROPES) type eye plaques loaded with I-125 (model 6711) seeds using GafChromic{sup ®} EBT3 films, in order to verify the dose distributions in the Plaque Simulator™ (PS) ophthalmic 3D treatment planning system. The brachytherapy module of RADCALC{sup ®} was used to independently check the dose distributions calculated by PS. Correction factors were derived from the measured data to be used in PS to account for the effect of the stainless steel ROPES plaque backing on the 3D dose distribution.Methods: Using GafChromic{sup ®} EBT3 films inserted in a specially designed Solid Water™ eye ball phantom, dose distributions were measured three-dimensionally both along and perpendicular to I-125 (model 6711) loaded ROPES eye plaque's central axis (CAX) with 2 mm depth increments. Each measurement was performed in full scatter conditions both with and without the stainless steel plaque backing attached to the eye plaque, to assess its effect on the dose distributions. Results were compared to the dose distributions calculated by Plaque Simulator™ and checked independently with RADCALC{sup ®}.Results: The EBT3 film measurements without the stainless steel backing were found to agree with PS and RADCALC{sup ®} to within 2% and 4%, respectively, on the plaque CAX. Also, RADCALC{sup ®} was found to agree with PS to within 2%. The CAX depth doses measured using EBT3 film with the stainless steel backing were observed to result in a 4% decrease relative to when the backing was not present. Within experimental uncertainty, the 4% decrease was found to be constant with depth and independent of plaque size. Using a constant dose correction factor of T= 0.96 in PS, where the calculated dose for the full water scattering medium is reduced by 4% in every voxel in the dose grid, the effect of the plaque backing was accurately

  10. Linearization of dose-response curve of the radiochromic film dosimetry system

    SciTech Connect

    Devic, Slobodan; Tomic, Nada; Aldelaijan, Saad; DeBlois, Francois; Seuntjens, Jan; Chan, Maria F.; Lewis, Dave

    2012-08-15

    Purpose: Despite numerous advantages of radiochromic film dosimeter (high spatial resolution, near tissue equivalence, low energy dependence) to measure a relative dose distribution with film, one needs to first measure an absolute dose (following previously established reference dosimetry protocol) and then convert measured absolute dose values into relative doses. In this work, we present result of our efforts to obtain a functional form that would linearize the inherently nonlinear dose-response curve of the radiochromic film dosimetry system. Methods: Functional form [{zeta}= (-1){center_dot}netOD{sup (2/3)}/ln(netOD)] was derived from calibration curves of various previously established radiochromic film dosimetry systems. In order to test the invariance of the proposed functional form with respect to the film model used we tested it with three different GAFCHROMIC Trade-Mark-Sign film models (EBT, EBT2, and EBT3) irradiated to various doses and scanned on a same scanner. For one of the film models (EBT2), we tested the invariance of the functional form to the scanner model used by scanning irradiated film pieces with three different flatbed scanner models (Epson V700, 1680, and 10000XL). To test our hypothesis that the proposed functional argument linearizes the response of the radiochromic film dosimetry system, verification tests have been performed in clinical applications: percent depth dose measurements, IMRT quality assurance (QA), and brachytherapy QA. Results: Obtained R{sup 2} values indicate that the choice of the functional form of the new argument appropriately linearizes the dose response of the radiochromic film dosimetry system we used. The linear behavior was insensitive to both film model and flatbed scanner model used. Measured PDD values using the green channel response of the GAFCHROMIC Trade-Mark-Sign EBT3 film model are well within {+-}2% window of the local relative dose value when compared to the tabulated Cobalt-60 data. It was also

  11. Evaluation of MVCT imaging dose levels during helical IGRT: comparison between ion chamber, TLD, and EBT3 films.

    PubMed

    Mege, Jean-Pierre; Wenzhao, Sun; Veres, Attila; Auzac, Guillaume; Diallo, Ibrahima; Lefkopoulos, Dimitri

    2016-01-01

    The purpose of this investigation was to evaluate the dose on megavoltage CT (MVCT) images required for tomotherapy. As imaging possibilities are often used before each treatment and usually used several times before the session, we tried to evaluate the dose delivered during the procedure. For each scanning mode (fine, normal, and coarse), we first established the relative variation of these doses according to different technical parameters (explored length, patient setup). These dose variations measured with the TomoPhant, also known as Cheese phantom, showed the expected variations (due to the variation of scattered radiation) of 15% according to the explored length and ± 5% according to the phantom setup (due to the variation of the point of measurement in the bore). In order to estimate patient doses, an anthropomorphic phantom was used for thermoluminescent and film dosimetry. The degree of agreement between the two methods was very satisfactory (the differences correspond to 5 mGy per imaging session) for the three sites studied (head & neck, thorax, and abdomen). These measurements allowed us to estimate the delivered dose of between 1 cGy and 4 cGy according to the site and imaging mode. Finally, we attempted to investigate a way to calculate this delivered dose in our patients from the study conducted on a cylindrical phantom and by taking into account data from the initial kV-CT scan. The results we obtained were close to our measurements, with discrepancies below 5 mGy per MVCT. PMID:26894346

  12. Experimental determination of the radial dose distribution in high gradient regions around {sup 192}Ir wires: Comparison of electron paramagnetic resonance imaging, films, and Monte Carlo simulations

    SciTech Connect

    Kolbun, N.; Leveque, Ph.; Abboud, F.; Bol, A.; Vynckier, S.; Gallez, B.

    2010-10-15

    Purpose: The experimental determination of doses at proximal distances from radioactive sources is difficult because of the steepness of the dose gradient. The goal of this study was to determine the relative radial dose distribution for a low dose rate {sup 192}Ir wire source using electron paramagnetic resonance imaging (EPRI) and to compare the results to those obtained using Gafchromic EBT film dosimetry and Monte Carlo (MC) simulations. Methods: Lithium formate and ammonium formate were chosen as the EPR dosimetric materials and were used to form cylindrical phantoms. The dose distribution of the stable radiation-induced free radicals in the lithium formate and ammonium formate phantoms was assessed by EPRI. EBT films were also inserted inside in ammonium formate phantoms for comparison. MC simulation was performed using the MCNP4C2 software code. Results: The radical signal in irradiated ammonium formate is contained in a single narrow EPR line, with an EPR peak-to-peak linewidth narrower than that of lithium formate ({approx}0.64 and 1.4 mT, respectively). The spatial resolution of EPR images was enhanced by a factor of 2.3 using ammonium formate compared to lithium formate because its linewidth is about 0.75 mT narrower than that of lithium formate. The EPRI results were consistent to within 1% with those of Gafchromic EBT films and MC simulations at distances from 1.0 to 2.9 mm. The radial dose values obtained by EPRI were about 4% lower at distances from 2.9 to 4.0 mm than those determined by MC simulation and EBT film dosimetry. Conclusions: Ammonium formate is a suitable material under certain conditions for use in brachytherapy dosimetry using EPRI. In this study, the authors demonstrated that the EPRI technique allows the estimation of the relative radial dose distribution at short distances for a {sup 192}Ir wire source.

  13. Calibration of GafChromic XR-RV3 radiochromic film for skin dose measurement using standardized x-ray spectra and a commercial flatbed scanner

    PubMed Central

    McCabe, Bradley P.; Speidel, Michael A.; Pike, Tina L.; Van Lysel, Michael S.

    2011-01-01

    Purpose: In this study, newly formulated XR-RV3 GafChromic® film was calibrated with National Institute of Standards and Technology (NIST) traceability for measurement of patient skin dose during fluoroscopically guided interventional procedures. Methods: The film was calibrated free-in-air to air kerma levels between 15 and 1100 cGy using four moderately filtered x-ray beam qualities (60, 80, 100, and 120 kVp). The calibration films were scanned with a commercial flatbed document scanner. Film reflective density-to-air kerma calibration curves were constructed for each beam quality, with both the orange and white sides facing the x-ray source. A method to correct for nonuniformity in scanner response (up to 25% depending on position) was developed to enable dose measurement with large films. The response of XR-RV3 film under patient backscattering conditions was examined using on-phantom film exposures and Monte Carlo simulations. Results: The response of XR-RV3 film to a given air kerma depended on kVp and film orientation. For a 200 cGy air kerma exposure with the orange side of the film facing the source, the film response increased by 20% from 60 to 120 kVp. At 500 cGy, the increase was 12%. When 500 cGy exposures were performed with the white side facing the x-ray source, the film response increased by 4.0% (60 kVp) to 9.9% (120 kVp) compared to the orange-facing orientation. On-phantom film measurements and Monte Carlo simulations show that using a NIST-traceable free-in-air calibration curve to determine air kerma in the presence of backscatter results in an error from 2% up to 8% depending on beam quality. The combined uncertainty in the air kerma measurement from the calibration curves and scanner nonuniformity correction was ±7.1% (95% C.I.). The film showed notable stability. Calibrations of film and scanner separated by 1 yr differed by 1.0%. Conclusions: XR-RV3 radiochromic film response to a given air kerma shows dependence on beam quality and film

  14. Dosimetry of very high energy electrons (VHEE) for radiotherapy applications: using radiochromic film measurements and Monte Carlo simulations

    NASA Astrophysics Data System (ADS)

    Subiel, A.; Moskvin, V.; Welsh, G. H.; Cipiccia, S.; Reboredo, D.; Evans, P.; Partridge, M.; DesRosiers, C.; Anania, M. P.; Cianchi, A.; Mostacci, A.; Chiadroni, E.; Di Giovenale, D.; Villa, F.; Pompili, R.; Ferrario, M.; Belleveglia, M.; Di Pirro, G.; Gatti, G.; Vaccarezza, C.; Seitz, B.; Isaac, R. C.; Brunetti, E.; Wiggins, S. M.; Ersfeld, B.; Islam, M. R.; Mendonca, M. S.; Sorensen, A.; Boyd, M.; Jaroszynski, D. A.

    2014-10-01

    Very high energy electrons (VHEE) in the range from 100-250 MeV have the potential of becoming an alternative modality in radiotherapy because of their improved dosimetry properties compared with MV photons from contemporary medical linear accelerators. Due to the need for accurate dosimetry of small field size VHEE beams we have performed dose measurements using EBT2 Gafchromic® film. Calibration of the film has been carried out for beams of two different energy ranges: 20 MeV and 165 MeV from conventional radio frequency linear accelerators. In addition, EBT2 film has been used for dose measurements with 135 MeV electron beams produced by a laser-plasma wakefield accelerator. The dose response measurements and percentage depth dose profiles have been compared with calculations carried out using the general-purpose FLUKA Monte Carlo (MC) radiation transport code. The impact of induced radioactivity on film response for VHEEs has been evaluated using the MC simulations. A neutron yield of the order of 10-5 neutrons cm-2 per incident electron has been estimated and induced activity due to radionuclide production is found to have a negligible effect on total dose deposition and film response. Neutron and proton contribution to the equivalent doses are negligible for VHEE. The study demonstrates that EBT2 Gafchromic film is a reliable dosimeter that can be used for dosimetry of VHEE. The results indicate an energy-independent response of the dosimeter for 20 MeV and 165 MeV electron beams and has been found to be suitable for dosimetry of VHEE.

  15. Dosimetry of very high energy electrons (VHEE) for radiotherapy applications: using radiochromic film measurements and Monte Carlo simulations.

    PubMed

    Subiel, A; Moskvin, V; Welsh, G H; Cipiccia, S; Reboredo, D; Evans, P; Partridge, M; DesRosiers, C; Anania, M P; Cianchi, A; Mostacci, A; Chiadroni, E; Di Giovenale, D; Villa, F; Pompili, R; Ferrario, M; Belleveglia, M; Di Pirro, G; Gatti, G; Vaccarezza, C; Seitz, B; Isaac, R C; Brunetti, E; Wiggins, S M; Ersfeld, B; Islam, M R; Mendonca, M S; Sorensen, A; Boyd, M; Jaroszynski, D A

    2014-10-01

    Very high energy electrons (VHEE) in the range from 100-250 MeV have the potential of becoming an alternative modality in radiotherapy because of their improved dosimetry properties compared with MV photons from contemporary medical linear accelerators. Due to the need for accurate dosimetry of small field size VHEE beams we have performed dose measurements using EBT2 Gafchromic® film. Calibration of the film has been carried out for beams of two different energy ranges: 20 MeV and 165 MeV from conventional radio frequency linear accelerators. In addition, EBT2 film has been used for dose measurements with 135 MeV electron beams produced by a laser-plasma wakefield accelerator. The dose response measurements and percentage depth dose profiles have been compared with calculations carried out using the general-purpose FLUKA Monte Carlo (MC) radiation transport code. The impact of induced radioactivity on film response for VHEEs has been evaluated using the MC simulations. A neutron yield of the order of 10(-5) neutrons cm(-2) per incident electron has been estimated and induced activity due to radionuclide production is found to have a negligible effect on total dose deposition and film response. Neutron and proton contribution to the equivalent doses are negligible for VHEE. The study demonstrates that EBT2 Gafchromic film is a reliable dosimeter that can be used for dosimetry of VHEE. The results indicate an energy-independent response of the dosimeter for 20 MeV and 165 MeV electron beams and has been found to be suitable for dosimetry of VHEE. PMID:25207591

  16. EBT reactor analysis

    SciTech Connect

    Uckan, N. A.; Jaeger, E. F.; Santoro, R. T.; Spong, D. A.; Uckan, T.; Owen, L. W.; Barnes, J. M.; McBride, J. B.

    1983-08-01

    This report summarizes the results of a recent ELMO Bumpy Torus (EBT) reactor study that includes ring and core plasma properties with consistent treatment of coupled ring-core stability criteria and power balance requirements. The principal finding is that constraints imposed by these coupling and other physics and technology considerations permit a broad operating window for reactor design optimization. Within this operating window, physics and engineering systems analysis and cost sensitivity studies indicate that reactors with <..beta../sub core/> approx. 6 to 10%, P approx. 1200 to 1700 MW(e), wall loading approx. 1.0 to 2.5 MW/m/sup 2/, and recirculating power fraction (including ring-sustaining power and all other reactors auxiliaries) approx. 10 to 15% are possible. A number of concept improvements are also proposed that are found to offer the potential for further improvement of the reactor size and parameters. These include, but are not limited to, the use of: (1) supplementary coils or noncircular mirror coils to improve magnetic geometry and reduce size, (2) energetic ion rings to improve ring power requirements, (3) positive potential to enhance confinement and reduce size, and (4) profile control to improve stability and overall fusion power density.

  17. Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies.

    PubMed

    Chiu-Tsao, Sou-Tung; Ho, Yunsil; Shankar, Ravi; Wang, Lin; Harrison, Louis B

    2005-11-01

    The purpose of this paper is to evaluate the energy dependence of the response of two new high sensitivity models of radiochromic films EBT and XR-QA. We determined the dose response curves of these films for four different radiation sources, namely, 6 MV photon beams (6 MVX), Ir-192, I-125, and Pd-103. The first type (EBT) is designed for intensity modulated radiation therapy (IMRT) dosimetry, and the second type (XR-QA) is designed for kilovoltage dosimetry. All films were scanned using red (665 nm) and green (520 nm) light sources in a charge-coupled device-based densitometer. The dose response curves [net optical density (NOD) versus dose] were plotted and compared for different radiation energies and light sources. Contrary to the early GAFCHROMIC film types (such as models XR, HS, MD55-2, and HD810), the net optical densities of both EBT and XR-QA were higher with a green (520 nm) than those with a red (665 nm) light source due to the different absorption spectrum of the new radiochromic emulsion. Both film types yield measurable optical densities for doses below 2 Gy. EBT film response is nearly independent of radiation energy, within the uncertainty of measurement. The NOD values of EBT film at 1 and 2 Gy are 0.13 and 0.25 for green, and 0.1 and 0.17 for red, respectively. In contrast, the XR-QA film sensitivity varies with radiation energy. The doses required to produce NOD of 0.5 are 6.9, 5.4, 0.7, and 0.9 Gy with green light and 19, 13, 1.7, and 1.5 Gy with red light, for 6 MVX, Ir-192, I -125, and Pd-103, respectively. EBT film was found to have minimal photon energy dependence of response for the energies tested and is suitable for dosimetry of radiation with a wide energy spectrum, including primary and scattered radiation. XR-QA film is promising for kilovoltage sources with a narrow energy spectra. The new high sensitivity radiochromic films are promising tools in radiation dosimetry. PMID:16370422

  18. Dosimetric response of radiochromic films to protons of low energies in the Bragg peak region

    NASA Astrophysics Data System (ADS)

    Battaglia, M. C.; Schardt, D.; Espino, J. M.; Gallardo, M. I.; Cortés-Giraldo, M. A.; Quesada, J. M.; Lallena, A. M.; Miras, H.; Guirado, D.

    2016-06-01

    One of the major advantages of proton or ion beams, applied in cancer treatment, is their excellent depth-dose profile exhibiting a low dose in the entrance channel and a distinct dose maximum (Bragg peak) near the end of range in tissue. In the region of the Bragg peak, where the protons or ions are almost stopped, experimental studies with low-energy particle beams and thin biological samples may contribute valuable information on the biological effectiveness in the stopping region. Such experiments, however, require beam optimization and special dosimetry techniques for determining the absolute dose and dose homogeneity for very thin biological samples. At the National Centre of Accelerators in Seville, one of the beam lines at the 3 MV Tandem Accelerator was equipped with a scattering device, a special parallel-plate ionization chamber with very thin electrode foils and target holders for cell cultures. In this work, we present the calibration in absolute dose of EBT3 films [Gafchromic radiotherapy films, http://www.ashland.com/products/gafchromic-radiotherapy-films] for proton energies in the region of the Bragg peak, where the linear energy transfer increases and becomes more significant for radiobiology studies, as well as the response of the EBT3 films for different proton energy values. To irradiate the films in the Bragg peak region, the energy of the beam was degraded passively, by interposing Mylar foils of variable thickness to place the Bragg peak inside the active layer of the film. The results obtained for the beam degraded in Mylar foils are compared with the dose calculated by means of the measurement of the beam fluence with an ionization chamber and the energy loss predicted by srim2008 code.

  19. A universal dose–response curve for radiochromic films

    SciTech Connect

    Martín-Viera Cueto, J. A. Parra Osorio, V.; Moreno Sáiz, C.; Navarro Guirado, F.; Casado Villalón, F. J.; Galán Montenegro, P.

    2015-01-15

    Purpose: This paper presents a model for dose–response curves of radiochromic films. It is based on a modified version of single-hit model to take into account the growth experienced by lithium salt of pentacosa-10,12-diynoic acid polymers after irradiation. Methods: Polymer growth in radiochromic films is a critical phenomenon that can be properly described by means of percolation theory to provide an appropriate distribution function for polymer sizes. Resulting functional form is a power function featuring a critical exponent and two adjustable parameters. Moreover, these parameters act as scaling factors setting a natural scale for sensitometric curves where the dependence on channel sensitivity is removed. A unique reduced response curve is then obtained from all the color channels describing film behavior independently of film dosimetry system. Results: Resulting functional form has been successfully tested in several sensitometric curves from different Gafchromic EBT models, providing excellent agreement with experimental data in a wide dose range up to about 40 Gy and low dose uncertainty. Conclusions: The model presented in this paper describes accurately the sensitometric curves of radiochromic films in wide dose ranges covering all typical ranges used in external radiotherapy. Resulting dose uncertainty is low enough to render a reasonably good performance in clinical applications. Due to cross-correlation, only one of the adjustable parameters is totally independent and characterizes film batches.

  20. INVESTIGATION OF A PLASMA MODE IN EBTS.

    SciTech Connect

    HERSHCOVITCH,A.

    2000-11-06

    A plasma related mode has been identified when EBTS operated with long trap length. The mode frequency scaling showed monotonic increased with confinement time. Initial scaling qualitatively suggested the mode to an electron beam driven ion cyclotron instability. However, a more quantitative evaluation is indicative of a drift mode. Nevertheless, the possibility of a structure mode, though unlikely, can not be completely excluded. The process of proper instability identification and stabilization is described.

  1. EBT-P proposed reference design report

    SciTech Connect

    Boch, A.L.

    1980-01-01

    This report describes the proposed reference design for the EBT-P proof-of-principle test device. The device described is a result of broad studies by many participating organizations from industry and from Department of Energy-sponsored fusion research groups, some working together and some in competitive studies, but all with the goal of defining a device at minimum cost and with maximum probability of meeting its goals. This design work is based upon advances in experimental and theoretical understanding of EBT achieved at Oak Ridge National Laboratory. The strategy adopted permits an initial test and validation of the key scaling properties of the ELMO Bumpy Torus concept, with a degree of built-in flexibility to extend the performance parameters toward the condition for containing a fusion reactor plasma. This will lead the way, then, to determination of a later power break-even demonstration and an eventual fusion reactor that can exploit the special high power-density and steady-state properties of the EBT concept.

  2. Development and characterization of a three-dimensional radiochromic film stack dosimeter for megavoltage photon beam dosimetry

    SciTech Connect

    McCaw, Travis J. Micka, John A.; DeWerd, Larry A.

    2014-05-15

    Purpose: Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry. Methods: A film stack dosimeter was developed using Gafchromic{sup ®} EBT2 films. The dosimeter consists of 22 films separated by 1 mm-thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed-dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage-depth-dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity-modulated radiation therapy (IMRT) procedure. Results: The absorbed-dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under-response when the

  3. Detailed modeling of microwave energy deposition in EBT devices

    SciTech Connect

    Batchelor, D.B.; Rasmussen, D.A.; Goldfinger, R.C.

    1983-08-01

    Ray-tracing studies have been combined with a simple wave power balance model to provide a complete, albeit approximate, description of microwave power deposition in the ELMO Bumpy Torus (EBT-I), EBT-Scale (EBT-S), and EBT Proof-of-Principle (EBT-P) devices. Electron cyclotron absorption of ordinary and extraordinary waves by the combined core plasma and relativistic annuli is calculated using a fully relativistic damping package developed for the RAYS geometrical optics code. The rays are traced in finite-beta bumpy cylinder plasma equilibria that are obtained from the Oak Ridge National Laboratory (ORNL) two-dimensional (2-D) equilibrium code. These results for direct, single-pass absorption are combined with results from a statistical model for the deposition of multiply reflected and mode-converted waves to obtain estimates of the power deposited in the core, surface, and annulus plasma components. Wave absorption by the annuli and by the core components at the fundamental and second harmonic resonances, reflection and Budden tunneling of the extraordinary mode at the right-hand cutoff, and conversion between ordinary and extraordinary modes upon wall reflection are the processes included in the power balance model. Experimental measurements of wave power flux on the cavity wall in EBT-S made with a simple microwave calorimeter are in good agreement with predictions of the model for a variety of operating configurations.

  4. Fractionated stereotactic radiotherapy: A method to evaluate geometric and dosimetric uncertainties using radiochromic films

    SciTech Connect

    Coscia, Gianluca; Vaccara, Elena; Corvisiero, Roberta; Cavazzani, Paolo; Ruggieri, Filippo Grillo; Taccini, Gianni

    2009-07-15

    In the authors' hospital, stereotactic radiotherapy treatments are performed with a Varian Clinac 600C equipped with a BrainLAB m3 micro-multileaf-collimator generally using the dynamic conformal arc technique. Patient immobilization during the treatment is achieved with a fixation mask supplied by BrainLAB, made with two reinforced thermoplastic sheets fitting the patient's head. With this work the authors propose a method to evaluate treatment geometric accuracy and, consequently, to determine the amount of the margin to keep in the CTV-PTV expansion during the treatment planning. The reproducibility of the isocenter position was tested by simulating a complete treatment on the anthropomorphic phantom Alderson Rando, inserting in between two phantom slices a high sensitivity Gafchromic EBT film, properly prepared and calibrated, and repeating several treatment sessions, each time removing the fixing mask and replacing the film inside the phantom. The comparison between the dose distributions measured on films and computed by TPS, after a precise image registration procedure performed by a commercial piece of software (FILMQA, 3cognition LLC (Division of ISP), Wayne, NJ), allowed the authors to measure the repositioning errors, obtaining about 0.5 mm in case of central spherical PTV and about 1.5 mm in case of peripheral irregular PTV. Moreover, an evaluation of the errors in the registration procedure was performed, giving negligible values with respect to the quantities to be measured. The above intrinsic two-dimensional estimate of treatment accuracy has to be increased for the error in the third dimension, but the 2 mm margin the authors generally use for the CTV-PTV expansion seems adequate anyway. Using the same EBT films, a dosimetric verification of the treatment planning system was done. Measured dose values are larger or smaller than the nominal ones depending on geometric irradiation conditions, but, in the authors' experimental conditions, always

  5. Fractionated stereotactic radiotherapy: a method to evaluate geometric and dosimetric uncertainties using radiochromic films.

    PubMed

    Coscia, Gianluca; Vaccara, Elena; Corvisiero, Roberta; Cavazzani, Paolo; Ruggieri, Filippo Grillo; Taccini, Gianni

    2009-07-01

    In the authors' hospital, stereotactic radiotherapy treatments are performed with a Varian Clinac 600C equipped with a BrainLAB m3 micro-multileaf-collimator generally using the dynamic conformal arc technique. Patient immobilization during the treatment is achieved with a fixation mask supplied by BrainLAB, made with two reinforced thermoplastic sheets fitting the patient's head. With this work the authors propose a method to evaluate treatment geometric accuracy and, consequently, to determine the amount of the margin to keep in the CTV-PTV expansion during the treatment planning. The reproducibility of the isocenter position was tested by simulating a complete treatment on the anthropomorphic phantom Alderson Rando, inserting in between two phantom slices a high sensitivity Gafchromic EBT film, properly prepared and calibrated, and repeating several treatment sessions, each time removing the fixing mask and replacing the film inside the phantom. The comparison between the dose distributions measured on films and computed by TPS, after a precise image registration procedure performed by a commercial piece of software (FILMQA, 3cognition LLC (Division of ISP), Wayne, NJ), allowed the authors to measure the repositioning errors, obtaining about 0.5 mm in case of central spherical PTV and about 1.5 mm in case of peripheral irregular PTV. Moreover, an evaluation of the errors in the registration procedure was performed, giving negligible values with respect to the quantities to be measured. The above intrinsic two-dimensional estimate of treatment accuracy has to be increased for the error in the third dimension, but the 2 mm margin the authors generally use for the CTV-PTV expansion seems adequate anyway. Using the same EBT films, a dosimetric verification of the treatment planning system was done. Measured dose values are larger or smaller than the nominal ones depending on geometric irradiation conditions, but, in the authors' experimental conditions, always

  6. A prototype, glassless densitometer traceable to primary optical standards for quantitative radiochromic film dosimetry

    SciTech Connect

    Rosen, B. S. Hammer, C. G.; Kunugi, K. A.; DeWerd, L. A.; Soares, C. G.

    2015-07-15

    Purpose: To evaluate a prototype densitometer traceable to primary optical standards and compare its performance to an EPSON Expression{sup ®} 10000XL flatbed scanner (the Epson) for quantitative radiochromic film (RCF) dosimetry. Methods: A prototype traceable laser densitometry system (LDS) was developed to mitigate common film scanning artifacts, such as positional scan dependence and high noise in low-dose regions, by performing point-based measurements of RCF suspended in free-space using coherent light. The LDS and the Epson optical absorbance scales were calibrated up to 3 AU, using reference materials calibrated at a primary standards laboratory and a scanner calibration factor (SCF). Calibrated optical density (OD) was determined for 96 Gafchromic{sup ®} EBT3 film segments before and after irradiation to one of 16 dose levels between 0 and 10 Gy, exposed to {sup 60}Co in a polymethyl-methacrylate (PMMA) phantom. The sensitivity was determined at each dose level and at two rotationally orthogonal readout orientations to obtain the sensitometric response of each RCF dosimetry system. LDS rotational scanning dependence was measured at nine angles between 0°and 180°, due to the expected interference between coherent light and polarizing EBT3 material. The response curves were fit to the analytic functions predicted by two physical response models: the two-parameter single-hit model and the four-parameter percolation model. Results: The LDS and the Epson absorbance measurements were linear to primary optical standards to within 0.2% and 0.3% up to 2 and 1 AU, respectively. At higher densities, the LDS had an over-response (2.5% at 3 AU) and the Epson an under-response (3.1% and 9.8% at 2 and 3 AU, respectively). The LDS and the Epson SCF over the applicable range were 0.968% ± 0.2% and 1.561% ± 0.3%, respectively. The positional scan dependence was evaluated on each digitizer and shown to be mitigated on the LDS, as compared to the Epson. Maximum EBT3

  7. EBT2 dosimetry of x-rays produced by the electron beam from a Plasma Focus for medical applications

    NASA Astrophysics Data System (ADS)

    Ceccolini, E.; Rocchi, F.; Mostacci, D.; Sumini, M.; Tartari, A.; Mariotti, F.

    2012-09-01

    The electron beam emitted from the back of Plasma Focus devices is being studied as a radiation source for intraoperative radiation therapy applications. A Plasma Focus device is being developed to this aim, to be utilized as an x-ray source. The electron beam is driven to impinge on 50 μm brass foil, where conversion x-rays are generated. Measurements with gafchromic film are performed to analyse the attenuation of the x-rays beam and to predict the dose given to the culture cell in radiobiological experiments to follow.

  8. EBT2 dosimetry of x-rays produced by the electron beam from a Plasma Focus for medical applications

    SciTech Connect

    Ceccolini, E.; Mostacci, D.; Sumini, M.; Rocchi, F.; Tartari, A.; Mariotti, F.

    2012-09-01

    The electron beam emitted from the back of Plasma Focus devices is being studied as a radiation source for intraoperative radiation therapy applications. A Plasma Focus device is being developed to this aim, to be utilized as an x-ray source. The electron beam is driven to impinge on 50 {mu}m brass foil, where conversion x-rays are generated. Measurements with gafchromic film are performed to analyse the attenuation of the x-rays beam and to predict the dose given to the culture cell in radiobiological experiments to follow.

  9. Elevation (typical EBT Coal Hopper Car) with scale. 3 Bay ...

    Library of Congress Historic Buildings Survey, Historic Engineering Record, Historic Landscapes Survey

    Elevation (typical EBT Coal Hopper Car) with scale. 3 Bay Steel Hopper Car with side extensions raising coal carrying capacity to 80,000 pounds. Note cars on either side lack extensions limiting their coal capacity to 70,000 pounds - East Broad Top Railroad & Coal Company, State Route 994, West of U.S. Route 522, Rockhill Furnace, Huntingdon County, PA

  10. SU-E-T-205: Improving Quality Assurance of HDR Brachytherapy: Verifying Agreement Between Planned and Delivered Dose Distributions Using DICOM RTDose and Advanced Film Dosimetry

    SciTech Connect

    Palmer, A L; Bradley, D A; Nisbet, A

    2014-06-01

    Purpose: HDR brachytherapy is undergoing significant development, and quality assurance (QA) checks must keep pace. Current recommendations do not adequately verify delivered against planned dose distributions: This is particularly relevant for new treatment planning system (TPS) calculation algorithms (non TG-43 based), and an era of significant patient-specific plan optimisation. Full system checks are desirable in modern QA recommendations, complementary to device-centric individual tests. We present a QA system incorporating TPS calculation, dose distribution export, HDR unit performance, and dose distribution measurement. Such an approach, more common in external beam radiotherapy, has not previously been reported in the literature for brachytherapy. Methods: Our QA method was tested at 24 UK brachytherapy centres. As a novel approach, we used the TPS DICOM RTDose file export to compare planned dose distribution with that measured using Gafchromic EBT3 films placed around clinical brachytherapy treatment applicators. Gamma analysis was used to compare the dose distributions. Dose difference and distance to agreement were determined at prescription Point A. Accurate film dosimetry was achieved using a glass compression plate at scanning to ensure physically-flat films, simultaneous scanning of known dose films with measurement films, and triple-channel dosimetric analysis. Results: The mean gamma pass rate of RTDose compared to film-measured dose distributions was 98.1% at 3%(local), 2 mm criteria. The mean dose difference, measured to planned, at Point A was -0.5% for plastic treatment applicators and -2.4% for metal applicators, due to shielding not accounted for in TPS. The mean distance to agreement was 0.6 mm. Conclusion: It is recommended to develop brachytherapy QA to include full-system verification of agreement between planned and delivered dose distributions. This is a novel approach for HDR brachytherapy QA. A methodology using advanced film

  11. Failure analysis of PB-1 (EBTS Be/Cu mockup)

    SciTech Connect

    Odegard, B.C. Jr.; Cadden, C.H.

    1996-11-01

    Failure analysis was done on PB-1 (series of Be tiles joined to Cu alloy) following a tile failure during a high heat flux experiment in EBTS (electron beam test system). This heat flux load simulated ambient conditions inside ITER; the Be tiles were bonded to the Cu alloy using low-temperature diffusion bonding, which is being considered for fabricating plasma facing components in ITER. Results showed differences between the EBTS failure and a failure during a room temperature tensile test. The latter occurred at the Cu-Be interface in an intermetallic phase formed by reaction of the two metals at the bonding temperature. Fracture strengths measured by these tests were over 300 MPa. The high heat flux specimens failed at the Cu-Cu diffusion bond. Fracture morphology in both cases was a mixed mode of dimple rupture and transgranular cleavage. Several explanations for this difference in failure mechanism are suggested.

  12. Particle confinement in EBT reactors with noncircular mirror coils

    SciTech Connect

    Owen, L.W.; Uckan, N.A.

    1983-01-01

    Methods of improving single particle confinement in the vacuum magnetic field of an ELMO Bumpy Torus (EBT) reactor have heretofore focused on enhancement of the effective magnetic aspect ratio through the addition of relatively low current supplementary coils to the basic EBT configuration of toroidally linked circular mirror coils. This method of aspect ratio enhancement is reviewed and compared to the use of noncircular, D-shaped mirror coils. A critical parameter in this evaluation is the required radial thickness delta of the blanket-shield assembly in the coil throat. Results indicate that D-coils represent an attractive alternative to the supplementary coil configurations if future neutronics calculations show that delta

  13. Poster — Thur Eve — 42: Radiochromic film calibration for low-energy seed brachytherapy dose measurement

    SciTech Connect

    Morrison, H; Menon, G; Sloboda, R

    2014-08-15

    The purpose of this study was to investigate the accuracy of radiochromic film calibration procedures used in external beam radiotherapy when applied to I-125 brachytherapy sources delivering higher doses, and to determine any necessary modifications to achieve similar accuracy in absolute dose measurements. GafChromic EBT3 film was used to measure radiation doses upwards of 35 Gy from 6 MV, 75 kVp and (∼28 keV) I-125 photon sources. A custom phantom was used for the I-125 irradiations to obtain a larger film area with nearly constant dose to reduce the effects of film heterogeneities on the optical density (OD) measurements. RGB transmission images were obtained with an Epson 10000XL flatbed scanner, and calibration curves relating OD and dose using a rational function were determined for each colour channel and at each energy using a non-linear least square minimization method. Differences found between the 6 MV calibration curve and those for the lower energy sources are large enough that 6 MV beams should not be used to calibrate film for low-energy sources. However, differences between the 75 kVp and I-125 calibration curves were quite small; indicating that 75 kVp is a good choice. Compared with I-125 irradiation, this gives the advantages of lower type B uncertainties and markedly reduced irradiation time. To obtain high accuracy calibration for the dose range up to 35 Gy, two-segment piece-wise fitting was required. This yielded absolute dose measurement accuracy above 1 Gy of ∼2% for 75 kVp and ∼5% for I-125 seed exposures.

  14. Advanced radiochromic film methodologies for quantitative dosimetry of small and nonstandard fields

    NASA Astrophysics Data System (ADS)

    Rosen, Benjamin S.

    Radiotherapy treatments with small and nonstandard fields are increasing in use as collimation and targeting become more advanced, which spare normal tissues while increasing tumor dose. However, dosimetry of small and nonstandard fields is more difficult than that of conventional fields due to loss of lateral charged-particle equilibrium, tight measurement setup requirements, source occlusion, and the volume-averaging effect of conventional dosimeters. This work aims to create new small and nonstandard field dosimetry protocols using radiochromic film (RCF) in conjunction with novel readout and analysis methodologies. It also is the intent of this work to develop an improved understanding of RCF structure and mechanics for its quantitative use in general applications. Conventional digitization techniques employ white-light, flatbed document scanners or scanning-laser densitometers which are not optimized for RCF dosimetry. A point-by-point precision laser densitometry system (LDS) was developed for this work to overcome the film-scanning artifacts associated with the use of conventional digitizers, such as positional scan dependence, off-axis light scatter, glass bed interference, and low signal-to-noise ratios. The LDS was shown to be optically traceable to national standards and to provide highly reproducible density measurements. Use of the LDS resulted in increased agreement between RCF dose measurements and the single-hit detector model of film response, facilitating traceable RCF calibrations based on calibrated physical quantities. GafchromicRTM EBT3 energy response to a variety of reference x-ray and gamma-ray beam qualities was also investigated. Conventional Monte Carlo methods are not capable of predicting film intrinsic energy response to arbitrary particle spectra. Therefore, a microdosimetric model was developed to simulate the underlying physics of the radiochromic mechanism and was shown to correctly predict the intrinsic response relative to a

  15. 7 CFR 274.8 - Functional and technical EBT system requirements.

    Code of Federal Regulations, 2011 CFR

    2011-01-01

    ... for the SNAP account; (ii) Comply with the requirements of 12 CFR part 205 (Regulation E) in addition... 7 Agriculture 4 2011-01-01 2011-01-01 false Functional and technical EBT system requirements. 274... PROGRAM BENEFITS § 274.8 Functional and technical EBT system requirements. (a) Functional...

  16. EBT-P gamma-ray shielding system

    SciTech Connect

    Gohar, Y.

    1981-12-01

    An elaborate study was carried out for the coil and biological shield of the ELMO Bumpy Torus proof-of-principle (EBT-P) device. A three-dimensional scoping study for the coil shield was performed for four different shielding options to define the heat load for each component and check the compliance with the design criterion of 10 watts maximum heat per coil from the gamma ray sources. Also, a detailed biological dose survey was performed which included: (a) the dose equivalent inside and outside the building, (b) the dose equivalent from the two mazes of the machine room, and (c) the skyshine contribution to the dose equivalent.

  17. Measurement of gold nanofilm dose enhancement using unlaminated radiochromic film

    SciTech Connect

    Rakowski, Joseph T. Snyder, Michael G.; Hillman, Yair; Laha, Suvra S.; Lawes, Gavin; Buczek, Matthew G.; Tucker, Mark A.; Liu, Fangchao; Mao, Guangzhao

    2015-10-15

    Purpose: Bombarding high-Z material with x-ray radiation releases Auger electrons and Coster–Kronig electrons, along with deeper penetrating fluorescent x-rays and photoelectrons. The Auger and Coster–Kronig electron penetration distance is on the order of nanometers to micrometers in water or tissue, creating a large dose enhancement accompanied by a RBE greater than 1 at the cellular level. The authors’ aim is to measure the gold nanofilm dose enhancement factor (DEF) at the cellular level with unlaminated radiochromic film via primary 50 kVp tungsten x-ray spectrum interaction, similar to an electronic brachytherapy spectrum. Methods: Unlaminated Gafchromic{sup ®} EBT2 film and Monte Carlo modeling were combined to derive DEF models. Gold film of thickness 23.1 ±  4.3 nm and surface roughness of 1.2 ± 0.2 nm was placed in contact with unlaminated radiochromic film in a downstream orientation and exposed to a 50 kVp tungsten bremsstrahlung, mean energy 19.2 keV. Film response correction factors were derived by Monte Carlo modeling of electron energy deposition in the film’s active layer, and by measuring film energy dependence from 4.5 keV to 50 kVp. Results: The measured DEF within a 13.6 μm thick water layer was 0.29 with a mean dose of 94 ± 9.4 cGy from Au emissions and 324 ± 32.4 cGy from the 50 kVp primary beam. Monte Carlo derived correction factors allowed determination of Au contributed dose in shallower depths at 0.25 μm intervals. Maximum DEF of 18.31 was found in the first 0.25 μm water depth. Conclusions: Dose enhancement from Au nanofilm can be measured at the cellular level using unlaminated radiochromic film. Complementing the measured dose value with Monte Carlo calculations allows estimation of dose enhancement at depth increments within the cellular range.

  18. Two-point model for electron transport in EBT

    SciTech Connect

    Chiu, S.C.; Guest, G.E.

    1980-01-01

    The electron transport in EBT is simulated by a two-point model corresponding to the central plasma and the edge. The central plasma is assumed to obey neoclassical collisionless transport. The edge plasma is assumed turbulent and modeled by Bohm diffusion. The steady-state temperatures and densities in both regions are obtained as functions of neutral influx and microwave power. It is found that as the neutral influx decreases and power increases, the edge density decreases while the core density increases. We conclude that if ring instability is responsible for the T-M mode transition, and if stability is correlated with cold electron density at the edge, it will depend sensitively on ambient gas pressure and microwave power.

  19. Analytical theory of interchange and compressional Alfven instabilities in EBT

    SciTech Connect

    Cheng, C.Z.; Tsang, K.T.

    1981-07-01

    The local stability of the EBT plasma is analyzed for the long wavelength perturbations in the frequency regime, ..omega.. approx. less than or equal to ..cap omega../sub i/(..cap omega../sub i/ is ion cyclotron frequency). In addition to the low frequency interchange instability, the plasma can be unstable to the compressional Alfven wave. Contrary to the previously obtained quadratic dispersion relation in ..omega.. for the interchange mode, our dispersion relations for both types of instabilities are cubic in ..omega... New stability boundaries are found, for the hot electron interchange mode, to relate to the enhanced compressibility of the core plasma in the presence of hot electrons. The compressional Alfven instability is driven due to the coupling of hot electron magnetic drifts and diamagnetic drift with the compressional Alfven wave. The stability conditions of these two types of instabilities are opposite to each other.

  20. SU-E-T-447: Electronic Brachytherapy (EBT) Treatment of Cervical Cancer - First Clinical Experience

    SciTech Connect

    Johnson, D; Johnson, M; Thompson, J; Ahmad, S; Chan, L; Hausen, H

    2014-06-01

    Purpose: To study the first trial patient in which an electronic brachytherapy (EBT) x-ray source is utilized for treatment of cervical cancer. Methods: During patient treatment, a miniaturized x-ray source was used in combination with a customized titanium tandem and ovoid applicator set. The semi-specialized source was modeled with formalisms outlined by AAMP Task Group 43. Multiple models were used to compensate for variable attenuation conditions as a function of source positions. Varian Brachyvision treatment planning software was utilized on CT data sets for dose calculations prior to treatment delivery. The dose was prescribed to “point A” as defined by American Brachytherapy society. Additional treatments plans were created from those clinically utilized in patient care and were recalculated for an existing Ir-192 source model. Dose volume histograms (DVH) and point dose calculations were compared between the modalities for the clinical condition present in patients treated with EBT. Results: Clinical treatment times, though longer than those typically experienced by Ir-192 users, were manageable. Instantaneous dose rates at personal positions within the treatment vault were lower than those measured during intra operative radiation therapy and breast EBT treatments. Due to lower average photon energy in EBT, dose gradients within the treatment plans were as expected steeper than those observed in Ir-192 based brachytherapy. DVH comparisons between Ir-192 and EBT treatments showed an expected decrease in the integral dose to normal tissues of interest for EBT. In comparing plans created for EBT delivery with those calculated for Ir-192, average dose values for EBT were more than 4%, 11%, and 9% lower at predefined bladder, rectum and “point B” positions, respectively. Conclusion: For the first time, we have demonstrated that the utilizing electronic brachytherapy system for tandem and ovoid based treatment of cancer of the cervix is feasible, and

  1. Application of the gamma evaluation method in Gamma Knife film dosimetry

    SciTech Connect

    Park, Jeong-Hoon; Han, Jung Ho; Kim, Chae-Yong; Oh, Chang Wan; Lee, Do-Heui; Suh, Tae-Suk; Gyu Kim, Dong; Chung, Hyun-Tai

    2011-10-15

    Purpose: Gamma Knife (GK) radiosurgery is a minimally invasive surgical technique for the treatment of intracranial lesions. To minimize neurological deficits, submillimeter accuracy is required during treatment delivery. In this paper, the delivery accuracy of GK radiosurgery was assessed with the gamma evaluation method using planning dose distribution and film measurement data. Methods: Single 4, 8, and 16 mm and composite shot plans were developed for evaluation using the GK Perfexion (PFX) treatment planning system (TPS). The planning dose distributions were exported as digital image communications in medicine - radiation therapy (DICOM RT) files using a new function of GK TPS. A maximum dose of 8 Gy was prescribed for four test plans. Irradiation was performed onto a spherical solid water phantom using Gafchromic EBT2 films in the axial and coronal planes. The exposed films were converted to absolute dose based on a 4th-order polynomial calibration curve determined using ten calibration films. The film measurement results and planning dose distributions were registered for further analysis in the same Leksell coordinate using in-house software. The gamma evaluation method was applied to two dose distributions with varying spatial tolerance (0.3-2.0 mm) and dosimetric tolerance (0.3-2.0%), to verify the accuracy of GK radiosurgery. The result of gamma evaluation was assessed using pass rate, dose gamma index histogram (DGH), and dose pass rate histogram (DPH). Results: The 20, 50, and 80% isodose lines found in film measurements were in close agreement with the planning isodose lines, for all dose levels. The comparison of diagonal line profiles across the axial plane yielded similar results. The gamma evaluation method resulted in high pass rates of >95% within the 50% isodose line for 0.5 mm/0.5% tolerance criteria, in both the axial and coronal planes. They satisfied 1.0 mm/1.0% criteria within the 20% isodose line. Our DGH and DPH also showed that low

  2. Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)

    NASA Astrophysics Data System (ADS)

    Purwaningsih, S.; Lubis, L. E.; Pawiro, S. A.; Soejoko, D. S.

    2016-03-01

    This research was aimed to check the patterns of dose profile on adult and pediatric head scan. We compared measurement result on dose profile along the z- axis rotation at peripheries and center phantom with a variety of pitch, i.e. 0.75, 1, 1.5 for adult and pediatric head protocol, keeping the rest of the scan parameters constant. Measurements were performed on homogeneous, cylindrical PMMA phantom with diameters of 16 and 10 cm using XR-QA2 Gafchromic film and TLD as dosimeters. The measurement result indicated a decrease in the dose about 50% and 47% for adult and pediatric head scan with the increase of pitch. For 0.75 value of pitch adult head scan, dose range for each position were (2.4 - 5.0) cGy, (3.1 - 5.3) cGy, (2.2 - 4.5) cGy, (2.8 - 5.3) cGy, and (3.3 - 5.6) cGy for position of center, 3, 6, 9 and 12 o'clock peripheral phantom position respectively. Dose profile for adult and pediatric head scan protocols has pattern curve with the maximum dose in the middle and tendency of symmetry near the edges, with different the plateau length along z- axis direction in accordance to the measurement position in the phantom.

  3. Vacuum system problems of EBT: a steady-state fusion experiment

    SciTech Connect

    Livesey, R.L.

    1981-01-01

    Many of the vacuum problems faced by EBT will soon be shared by other plasma devices as high-power microwave systems and long pulse lengths become more common. The solutions used on EBT (such as the raised lip with elastomer seal) are not unique; however, experience has shown that microwave-compatible designs must be carefully thought out. All details of the vacuum must be carefully thought out. All details of the vacuum must be carefully screened in advance to insure that microwaves do not leak into pumps or diagnostics where they can cause major damage. Sputter coating, which even now is noticeably present in most pulsed plasma systems, becomes much worse as systems approach steady state. And finally, radiation degradation of components which is presently a minor problem will become significant on high-power microwave-fed devices, such as EBT-P.

  4. Total skin electron therapy (TSET): A reimplementation using radiochromic films and IAEA TRS-398 code of practice

    SciTech Connect

    Schiapparelli, P.; Zefiro, D.; Massone, F.; Taccini, G.

    2010-07-15

    Purpose: The aim of this work is to present an updated implementation of total skin electron therapy (TSET) using IAEA TRS-398 code of practice for absolute dosimetry and taking advantage of the use of radiochromic films. The optimization of quality control tests is also included. Methods: A Varian 2100 C/D linear accelerator equipped with the special procedure HDTSe{sup -} (high dose rate total skin electron mode, E=6 MeV) was employed to perform TSET irradiations using the modified Stanford technique. The commissioning was performed following the AAPM report 23 recommendations. In particular, for dual-field beams irradiation, the optimal tilt angle was investigated and the dose distribution in the treatment plane was measured. For a complete six dual-field beams irradiation, the treatment skin dose on the surface of a cylindrical phantom was evaluated by radiochromic films and the B factor which relates the single dual-field skin dose to the six dual-field skin dose was assessed. Since the TRS-398 reference conditions do not meet the requirements of TSET absolute dosimetry, GafChromic EBT films were also employed to check and validate the application of the protocol. Simplified procedures were studied to verify beam constancy in PMMA phantoms without the more difficult setup of total skin irradiation. Results: The optimized geometrical setup for dual-field beams was: Tilt angle={+-}19 deg., SSD=353 cm, and the beam degrader (200x100x1 cm{sup 3}) placed at 320 cm from the source. As regards to dose homogeneity in the treatment plane, for dual-field beams irradiation, the mean relative dose value was 97%{+-}5% (normalizing to 100% at the calibration point level). For six dual-field beams irradiation, the multiplication factor B was 2.63. In addition, beam quality, dose rate, and bremsstrahlung contribution were also suitable for TSET treatments. The TRS-398 code of practice was used for TSET dosimetry, as dose measurements performed by ionization chamber and

  5. 7 CFR 274.8 - Functional and technical EBT system requirements.

    Code of Federal Regulations, 2014 CFR

    2014-01-01

    ... standard response time. (2) System availability and reliability. (i) The EBT system central computer shall... maintenance. The total system, including the system's central computer, any network or intermediate processing... prescribed herein prior to permitting the interface with the central computer system. (ii) The system......

  6. 7 CFR 274.8 - Functional and technical EBT system requirements.

    Code of Federal Regulations, 2012 CFR

    2012-01-01

    ... standard response time. (2) System availability and reliability. (i) The EBT system central computer shall... maintenance. The total system, including the system's central computer, any network or intermediate processing... prescribed herein prior to permitting the interface with the central computer system. (ii) The system......

  7. 7 CFR 274.8 - Functional and technical EBT system requirements.

    Code of Federal Regulations, 2013 CFR

    2013-01-01

    ... standard response time. (2) System availability and reliability. (i) The EBT system central computer shall... maintenance. The total system, including the system's central computer, any network or intermediate processing... prescribed herein prior to permitting the interface with the central computer system. (ii) The system......

  8. 49 CFR 40.233 - What are the requirements for proper use and care of EBTs?

    Code of Federal Regulations, 2010 CFR

    2010-10-01

    ... 49 Transportation 1 2010-10-01 2010-10-01 false What are the requirements for proper use and care... Supplies Used in Alcohol Testing § 40.233 What are the requirements for proper use and care of EBTs? (a) As... must take into account factors like frequency of use, environmental conditions (e.g.,...

  9. Comparative measurement of the neutral density and particle confinement time in EBT

    SciTech Connect

    Glowienka, John C.; Richards, Roger K.

    1985-11-01

    The neutral density and particle confinement time in the ELMO Bumpy Torus-Scale Experiment (EBT-S) have been determined by two different techniques. These involve a spectroscopic measurement of molecular and atomic hydrogen emissions and a time-decay measurement of a fast-ion population using a diagnostic neutral beam. The results from both diagnostics exhibit identical trends for either estimate, although the absolute values differ by a factor of 2 to 3. The observed variations with fill gas pressure and microwave power from either technique are consistent with measurements of electron density and temperature. In this paper, the measurement techniques are discussed, and the results are compared in the context of consistency with independently observed plasma behavior.

  10. EBT Fidelity Trajectories Across Training Cohorts Using the Interagency Collaborative Team Strategy

    PubMed Central

    Hecht, Debra; Aarons, Greg; Fettes, Danielle; Hurlburt, Michael; Ledesma, Karla

    2015-01-01

    The Interdisciplinary Collaborative Team (ICT) strategy uses front-line providers as adaptation, training and quality control agents for multi-agency EBT implementation. This study tests whether an ICT transmits fidelity to subsequent provider cohorts. SafeCare was implemented by home visitors from multiple community-based agencies contracting with child welfare. Client-reported fidelity trajectories for 5,769 visits, 957 clients and 45 providers were compared using three-level growth models. Provider cohorts trained and live-coached by the ICT attained benchmark fidelity after 12 weeks, and this was sustained. Hispanic clients reported high cultural competency, supporting a cultural adaptation crafted by the ICT. PMID:25586878

  11. Validation of fast Monte Carlo dose calculation in small animal radiotherapy with EBT3 radiochromic films.

    PubMed

    Noblet, C; Chiavassa, S; Smekens, F; Sarrut, D; Passal, V; Suhard, J; Lisbona, A; Paris, F; Delpon, G

    2016-05-01

    In preclinical studies, the absorbed dose calculation accuracy in small animals is fundamental to reliably investigate and understand observed biological effects. This work investigated the use of the split exponential track length estimator (seTLE), a new kerma based Monte Carlo dose calculation method for preclinical radiotherapy using a small animal precision micro irradiator, the X-RAD 225Cx. Monte Carlo modelling of the irradiator with GATE/GEANT4 was extensively evaluated by comparing measurements and simulations for half-value layer, percent depth dose, off-axis profiles and output factors in water and water-equivalent material for seven circular fields, from 20 mm down to 1 mm in diameter. Simulated and measured dose distributions in cylinders of water obtained for a 360° arc were also compared using dose, distance-to-agreement and gamma-index maps. Simulations and measurements agreed within 3% for all static beam configurations, with uncertainties estimated to 1% for the simulation and 3% for the measurements. Distance-to-agreement accuracy was better to 0.14 mm. For the arc irradiations, gamma-index maps of 2D dose distributions showed that the success rate was higher than 98%, except for the 0.1 cm collimator (92%). Using the seTLE method, MC simulations compute 3D dose distributions within minutes for realistic beam configurations with a clinically acceptable accuracy for beam diameter as small as 1 mm. PMID:27055114

  12. Validation of fast Monte Carlo dose calculation in small animal radiotherapy with EBT3 radiochromic films

    NASA Astrophysics Data System (ADS)

    Noblet, C.; Chiavassa, S.; Smekens, F.; Sarrut, D.; Passal, V.; Suhard, J.; Lisbona, A.; Paris, F.; Delpon, G.

    2016-05-01

    In preclinical studies, the absorbed dose calculation accuracy in small animals is fundamental to reliably investigate and understand observed biological effects. This work investigated the use of the split exponential track length estimator (seTLE), a new kerma based Monte Carlo dose calculation method for preclinical radiotherapy using a small animal precision micro irradiator, the X-RAD 225Cx. Monte Carlo modelling of the irradiator with GATE/GEANT4 was extensively evaluated by comparing measurements and simulations for half-value layer, percent depth dose, off-axis profiles and output factors in water and water-equivalent material for seven circular fields, from 20 mm down to 1 mm in diameter. Simulated and measured dose distributions in cylinders of water obtained for a 360° arc were also compared using dose, distance-to-agreement and gamma-index maps. Simulations and measurements agreed within 3% for all static beam configurations, with uncertainties estimated to 1% for the simulation and 3% for the measurements. Distance-to-agreement accuracy was better to 0.14 mm. For the arc irradiations, gamma-index maps of 2D dose distributions showed that the success rate was higher than 98%, except for the 0.1 cm collimator (92%). Using the seTLE method, MC simulations compute 3D dose distributions within minutes for realistic beam configurations with a clinically acceptable accuracy for beam diameter as small as 1 mm.

  13. Research support for plasma diagnostics on Elmo Bumpy Torus. Observations of EBT ring geometry by means of diamagnetic field measurements using a compensated ferromagnetic resonance magnetometer, and a preliminary numerical study of EBT flux penetration dynamics

    SciTech Connect

    Carpenter, K.H.; Booker, R.H.

    1982-10-01

    A preliminary numerical study of flux penetration dynamics has been made assuming a bumpy cylinder geometry for the EBT vacuum vessel. Estimates made by this study show that the diamagnetic signal from the rings should penetrate the vacuum vessel in a time on the order of 100 mS when the plasma is changed. This penetration time appears qualitatively reasonable.

  14. SU-E-T-37: An Optical Investigation Into the Polarization and Scattering Effects Underlying the Artifacts of Radiochromic Film Dosimetry with Commercial Flatbed Scanners

    SciTech Connect

    Schoenfeld, A; Poppinga, D; Poppe, B; Harder, D; Doerner, K

    2014-06-01

    Purpose: This study aims to investigate the optical properties of radiochromic EBT3 films on exposure to polarized incident light. Methods: An optical table setup was used to investigate the properties of exposed and unexposed EBT3 films. The films were placed with their long side horizontally and illuminated with polarized incident white light. The polarization of light with the electrical vector pointing vertically is referred to as 0°, accordingly horizontal orientation corresponds to 90°. The light transmission was measured depending on the polarization angle of the incident light and the polarization of a polarizer in front of the detector. Secondly, the scattering properties of exposed and unexposed films were measured by placing a plane convex lens behind the films and a screen in its focal plane. Thereby, the distribution of the scattering angles appears as an intensity map on the screen. The distributions of scattering angles caused by EBT3 films and by neutral density filters were compared. Results: EBT3 films show a strong dependence of the light transmission on the polarization of the incident light. With both polarizers parallel, a peak transmission was found at 90° orientation of the polarizers. With the rear polarizer at right angles with the front polarizer, peak transmissions were found at front polarizer orientations 45° and 135°. The scattering appears to be anisotropic with a preference direction parallel to the long side of the film. The portion of scattered light and the half value scattering angle both increase with the dose on the film. Conclusion: EBT3 films show dose dependent changes in polarized light transmission and anisotropic light scattering. These effects impair the light absorption measurements on exposed films performed with commercial flatbed scanners and are causing the well-known artifacts of radiochromic film dosimetry with flatbed scanners, the “orientation effect” and the “parabola effect”.

  15. The artefacts of radiochromic film dosimetry with flatbed scanners and their causation by light scattering from radiation-induced polymers.

    PubMed

    Schoenfeld, Andreas A; Poppinga, Daniela; Harder, Dietrich; Doerner, Karl-Joachim; Poppe, Bjoern

    2014-07-01

    Optical experiments and theoretical considerations have been undertaken in order to understand the causes of the 'orientation effect' and the 'parabola effect', the artefacts impairing the desired light absorption measurement on radiochromic EBT3 films with flatbed scanners. EBT3 films exposed to doses up to 20.9 Gy were scanned with an Epson Expression 10000XL flatbed scanner in landscape and portrait orientation. The horizontally and vertically polarized light components of the scanner were determined, and another Epson Expression 10000XL flatbed scanner was disassembled to examine its optical components. The optical properties of exposed and unexposed EBT3 films were studied with incident polarized and unpolarized white light, and the transmitted red light was investigated for its polarization and scattering properties including the distribution of the scattering angles. Neutral density filters were studied for comparison. Guidance was sought from the theory of light scattering from rod-like macromolecular structures. The drastic dose-dependent variation of the transmitted total light current as function of the orientation of front and rear polarizers, interpreted by light scattering theory, shows that the radiation-induced polymerization of the monomers of EBT3 films produces light scattering oscillators preferably polarized at right angles with the coating direction of the film. The directional distribution of the scattered light is partly anisotropic, with a preferred scattering plane at right angles with the coating direction, indicating light scattering from stacks of coherently vibrating oscillators piled up along the monomer crystals. The polyester carrier film also participates in these effects. The 'orientation' and 'parabola' artefacts due to flatbed scanning of radiochromic films can be explained by the interaction of the polarization-dependent and anisotropic light scattering from exposed and unexposed EBT3 films with the quantitative difference

  16. NOTE: Use of a blood glucose meter for radiochromic film analysis in blood irradiation

    NASA Astrophysics Data System (ADS)

    Cheung, Tsang; Butson, Martin J.; Yu, Peter K. N.

    2002-10-01

    The use of a diabetic blood glucose meter for radiochromic film dosimetry in blood irradiation using x-ray beams on a medical linear accelerator has been investigated. The glucose meter provides optical density analysis in the visible and infrared region using a reflectance measurement technique. By comparing the 'blood sugar' level output with standard calibration gafchromic films a calibration curve is produced for quantitative analysis. Results show that a reproducible dose to meter output curve can be fitted using a second order polynomial function and that blood irradiation doses in vitro were measured to within 7.9% mean error (as compared to ionization chamber results) using the blood glucose meter. This level of accuracy falls below that measured with a standard densitometer (4.3%) however, results show that the blood glucose meter, which would be available in any haematology department, produces an adequate measure of gafchromic film optical density for blood irradiation dosimetry.

  17. On multichannel film dosimetry with channel-independent perturbations

    SciTech Connect

    Méndez, I. Peterlin, P.; Hudej, R.; Strojnik, A.; Casar, B.

    2014-01-15

    Purpose: Different multichannel methods for film dosimetry have been proposed in the literature. Two of them are the weighted mean method and the method put forth byMicke et al. [“Multichannel film dosimetry with nonuniformity correction,” Med. Phys. 38, 2523–2534 (2011)] and Mayer et al. [“Enhanced dosimetry procedures and assessment for EBT2 radiochromic film,” Med. Phys. 39, 2147–2155 (2012)]. The purpose of this work was to compare their results and to develop a generalized channel-independent perturbations framework in which both methods enter as special cases. Methods: Four models of channel-independent perturbations were compared: weighted mean, Micke–Mayer method, uniform distribution, and truncated normal distribution. A closed-form formula to calculate film doses and the associated type B uncertainty for all four models was deduced. To evaluate the models, film dose distributions were compared with planned and measured dose distributions. At the same time, several elements of the dosimetry process were compared: film type EBT2 versus EBT3, different waiting-time windows, reflection mode versus transmission mode scanning, and planned versus measured dose distribution for film calibration and for γ-index analysis. The methods and the models described in this study are publicly accessible through IRISEU. Alpha 1.1 ( http://www.iriseu.com ). IRISEU. is a cloud computing web application for calibration and dosimetry of radiochromic films. Results: The truncated normal distribution model provided the best agreement between film and reference doses, both for calibration and γ-index verification, and proved itself superior to both the weighted mean model, which neglects correlations between the channels, and the Micke–Mayer model, whose accuracy depends on the properties of the sensitometric curves. With respect to the selection of dosimetry protocol, no significant differences were found between transmission and reflection mode scanning

  18. Measurements of proton energy spectra using a radiochromic film stack

    NASA Astrophysics Data System (ADS)

    Filkins, T. M.; Steidle, Jessica; Ellison, D. M.; Steidle, Jeffrey; Freeman, C. G.; Padalino, S. J.; Fiksel, G.; Regan, S. P.; Sangster, T. C.

    2014-10-01

    The energy spectrum of protons accelerated from the rear-side of a thin foil illuminated with ultra-intense laser light from the OMEGA EP laser system at the University of Rochester's Laboratory for Laser Energetics (LLE) was measured using a stack of radiochromic film (RCF). The film stack consisted of four layers of Gafchromic HD-V2 film and four layers of Gafchromic MD-V2-55 film. Aluminum foils of various thicknesses were placed between each piece of RCF in the stack. This arrangement allowed protons with energies of 30 MeV to reach the back layer of RCF in the stack. The stack was placed in the detector plane of a Thomson parabola ion energy (TPIE) spectrometer. Each piece of film in the stack was scanned using a commercially available flat-bed scanner (Epson 10000XL). The resulting optical density was converted into proton fluence using an absolute calibration of the RCF obtained at the SUNY Geneseo 1.7 MV Pelletron accelerator laboratory. In these calibration measurements, the sensitivity of the radiochromic film was measured using monoenergetic protons produced by the accelerator. Details of the analysis procedure and the resulting proton energy spectra will be presented. Funded in part by a grant from the DOE through the Laboratory for Laser Energetics.

  19. Evaluation of material heterogeneity dosimetric effects using radiochromic film for COMS eye plaques loaded with {sup 125}I seeds (model I25.S16)

    SciTech Connect

    Acar, Hilal; Chiu-Tsao, Sou-Tung; Oezbay, Ismail; Kemikler, Goenuel; Tuncer, Samuray

    2013-01-15

    Purpose: (1) To measure absolute dose distributions in eye phantom for COMS eye plaques with {sup 125}I seeds (model I25.S16) using radiochromic EBT film dosimetry. (2) To determine the dose correction function for calculations involving the TG-43 formalism to account for the presence of the COMS eye plaque using Monte Carlo (MC) method specific to this seed model. (3) To test the heterogeneous dose calculation accuracy of the new version of Plaque Simulator (v5.3.9) against the EBT film data for this seed model. Methods: Using EBT film, absolute doses were measured for {sup 125}I seeds (model I25.S16) in COMS eye plaques (1) along the plaque's central axis for (a) uniformly loaded plaques (14-20 mm in diameter) and (b) a 20 mm plaque with single seed, and (2) in off-axis direction at depths of 5 and 12 mm for all four plaque sizes. The EBT film calibration was performed at {sup 125}I photon energy. MC calculations using MCNP5 code for a single seed at the center of a 20 mm plaque in homogeneous water and polystyrene medium were performed. The heterogeneity dose correction function was determined from the MC calculations. These function values at various depths were entered into PS software (v5.3.9) to calculate the heterogeneous dose distributions for the uniformly loaded plaques (of all four sizes). The dose distributions with homogeneous water assumptions were also calculated using PS for comparison. The EBT film measured absolute dose rate values (film) were compared with those calculated using PS with homogeneous assumption (PS Homo) and heterogeneity correction (PS Hetero). The values of dose ratio (film/PS Homo) and (film/PS Hetero) were obtained. Results: The central axis depth dose rate values for a single seed in 20 mm plaque measured using EBT film and calculated with MCNP5 code (both in ploystyrene phantom) were compared, and agreement within 9% was found. The dose ratio (film/PS Homo) values were substantially lower than unity (mostly between 0.8 and 0

  20. Improving analysis of radiochromic films

    NASA Astrophysics Data System (ADS)

    Baptista Neto, A. T.; Meira-Belo, L. C.; Faria, L. O.

    2014-02-01

    An appropriate radiochromic film should respond uniformly throughout its surface after exposure to a uniform radiation field. The evaluation of radiochromic films may be carried out, for example, by measuring color intensities in film-based digitized images. Fluctuations in color intensity may be caused by many different factors such as optical structure of the film's active layer, defects in structure of the film, scratches and external agents, such as dust. The use of high spatial resolution during film scanning should also increase microscopic uniformity. Since the average is strongly influenced by extreme values, the use of other statistical tools, for which this problem becomes inconspicuous, optimizes the application of higher spatial resolution as well as reduces standard deviations. This paper compares the calibration curves of the XR-QA2 Gafchromic® radiochromic film based on three different methods: the average of all color intensity readings, the median of these same readings and the average of readings that fall between the first and third quartiles. Results indicate that a higher spatial resolution may be adopted whenever the calibration curve is based on tools less influenced by extreme values such as those generated by the factors mentioned above.

  1. Spectrophotometric determination of β-adrenergic antagonists drugs via ion-pair complex formation using MO and EBT

    NASA Astrophysics Data System (ADS)

    El-Didamony, A. M.; Shehata, A. M.

    2014-09-01

    Two simple, rapid and sensitive spectrophotometric methods have been proposed for the assay of bisoprolol fumarate (BSF), propranolol hydrochloride (PRH), and timolol maleate (TIM) either in bulk or in pharmaceutical formulations. The methods are based on the reaction of the selected drugs with methyl orange (MO) and eriochrome black T in acidic buffers, after extracting in dichloromethane and measured quantitatively with maximum absorption at 428 and 518 nm for MO and EBT, respectively. The analytical parameters and their effects on the reported systems are investigated. The extracts are intensely colored and very stable at room temperature. The calibration graphs were linear over the concentration range of 0.8-6.4, 0.4-3.6, 0.8-5.6 μg/mL for BSF, PRH, and TIM, respectively, with MO and 0.8-6.4, 0.4-3.2, and 0.8-8.0 μg/mL for BSF, PRH, and TIM, respectively, with EBT. The stoichiometry of the complexes was found to be 1 : 1 in all cases. The proposed methods were successfully extended to pharmaceutical preparations. Excipients used as additive in commercial formulations did not interfere in the analysis. The proposed methods can be recommended for quality control and routine analysis where time, cost effectiveness and high specificity of analytical technique are of great importance.

  2. Radiochromic dye film studies for brachytherapy applications.

    PubMed

    Martínez-Dávalos, A; Rodríguez-Villafuerte, M; Díaz-Perches, R; Arzamendi-Pérez, S

    2002-01-01

    Commercial radiochromic dye films have been used in recent years to quantify absorbed dose in several medical applications. In this study we present the characterisation of the GafChromic MD-55-2 dye film, a double sensitive layer film suitable for photon irradiation in brachytherapy applications. Dose measurements were carried out with a low dose rate 137Cs brachytherapy source, which produces very steep dose gradients in its vicinity, and therefore requires the capability of producing high spatial resolution isodose curves. Quantification of the dose rate in water per unit air kerma strength was obtained using a high-resolution transmission commercial scanner (Agfa DuoScan T1200 with the capability of digitising up to 600 x 1200 pixels per inch using 36 bits per pixel, together with optical density measurements. The Monte Carlo calculations and experimental measurements compared well in the 0-50 Gy dose interval used in this study. PMID:12382798

  3. Comparative Analysis of Different Measurement Techniques for MLC Characterization: Preliminary Results

    SciTech Connect

    Larraga-Gutierrez, J. M.; Ballesteros-Zebadua, P.; Garcia-Garduno, O. A.; Martinez-Davalos, A.; Rodriguez-Villafuerte, M.; Moreno-Jimenez, S.; Celis, M. A.

    2008-08-11

    Radiation transmission, leakage and beam penumbra are essential dosimetric parameters related to the commissioning of a multileaf collimation system. This work shows a comparative analysis of commonly used film detectors: X-OMAT V2 and EDR2 radiographic films, and GafChromic EBT registered radiochromic film. The results show that X-OMAT over-estimates radiation leakage and 80-20% beam penumbra. However, according to the reference values reported by the manufacturer for these dosimetric parameters, all three films are adequate for MLC dosimetric characterization, but special care must be taken when X-OMAT V2 film is used due to its low energy photon dependence.

  4. 49 CFR 40.243 - What is the procedure for an alcohol screening test using an EBT or non-evidential breath ASD?

    Code of Federal Regulations, 2013 CFR

    2013-10-01

    ... test using an EBT or non-evidential breath ASD? 40.243 Section 40.243 Transportation Office of the...-evidential breath ASD? As the BAT or STT, you must take the following steps: (a) Select, or allow the... breath has been obtained. (d) Show the employee the displayed test result. (e) If the device is one...

  5. 49 CFR 40.243 - What is the procedure for an alcohol screening test using an EBT or non-evidential breath ASD?

    Code of Federal Regulations, 2011 CFR

    2011-10-01

    ... test using an EBT or non-evidential breath ASD? 40.243 Section 40.243 Transportation Office of the...-evidential breath ASD? As the BAT or STT, you must take the following steps: (a) Select, or allow the... breath has been obtained. (d) Show the employee the displayed test result. (e) If the device is one...

  6. 49 CFR 40.243 - What is the procedure for an alcohol screening test using an EBT or non-evidential breath ASD?

    Code of Federal Regulations, 2014 CFR

    2014-10-01

    ... test using an EBT or non-evidential breath ASD? 40.243 Section 40.243 Transportation Office of the...-evidential breath ASD? As the BAT or STT, you must take the following steps: (a) Select, or allow the... breath has been obtained. (d) Show the employee the displayed test result. (e) If the device is one...

  7. 49 CFR 40.243 - What is the procedure for an alcohol screening test using an EBT or non-evidential breath ASD?

    Code of Federal Regulations, 2010 CFR

    2010-10-01

    ... test using an EBT or non-evidential breath ASD? 40.243 Section 40.243 Transportation Office of the...-evidential breath ASD? As the BAT or STT, you must take the following steps: (a) Select, or allow the... breath has been obtained. (d) Show the employee the displayed test result. (e) If the device is one...

  8. 49 CFR 40.243 - What is the procedure for an alcohol screening test using an EBT or non-evidential breath ASD?

    Code of Federal Regulations, 2012 CFR

    2012-10-01

    ... test using an EBT or non-evidential breath ASD? 40.243 Section 40.243 Transportation Office of the...-evidential breath ASD? As the BAT or STT, you must take the following steps: (a) Select, or allow the... breath has been obtained. (d) Show the employee the displayed test result. (e) If the device is one...

  9. Correcting scan-to-scan response variability for a radiochromic film-based reference dosimetry system

    SciTech Connect

    Lewis, David; Devic, Slobodan

    2015-10-15

    Purpose: In radiochromic film dosimetry systems, measurements are usually obtained from film images acquired on a CCD-based flatbed scanner. The authors investigated factors affecting scan-to-scan response variability leading to increased dose measurement uncertainty. Methods: The authors used flatbed document scanners to repetitively scan EBT3 radiochromic films exposed to doses 0–1000 cGy, together with three neutral density filters and three blue optical filters. Scanning was performed under two conditions: scanner lid closed and scanner lid opened/closed between scans. The authors also placed a scanner in a cold room at 9 °C and later in a room at 22 °C and scanned EBT3 films to explore temperature effects. Finally, the authors investigated the effect of altering the distance between the film and the scanner’s light source. Results: Using a measurement protocol to isolate the contribution of the CCD and electronic circuitry of the scanners, the authors found that the standard deviation of response measurements for the EBT3 film model was about 0.17% for one scanner and 0.09% for the second. When the lid of the first scanner was opened and closed between scans, the average scan-to-scan difference of responses increased from 0.12% to 0.27%. Increasing the sample temperature during scanning changed the RGB response values by about −0.17, −0.14, and −0.05%/°C, respectively. Reducing the film-to-light source distance increased the RBG response values about 1.1, 1.3, and 1.4%/mm, respectively. The authors observed that films and film samples were often not flat with some areas up to 8 mm away from the scanner’s glass window. Conclusions: In the absence of measures to deal with the response irregularities, each factor the authors investigated could lead to dose uncertainty >2%. Those factors related to the film-to-light source distance could be particularly impactful since the authors observed many instances where the curl of film samples had the

  10. Examination of the relevance of using radiochromic films in measuring entrance skin dose distribution in conventional digital mammography.

    PubMed

    Soliman, K; Bakkari, M

    2015-07-01

    Based on manufacturer specifications, radiochromic films are sensitive enough to be used for dosimetry in digital mammography (DM). The aim of this work was to study the feasibility of measuring entrance surface dose (ESD) distribution using Gafchromic XR-QA2 films. The films were irradiated following a standard clinical two-view screening mammography protocol using a full-field digital mammography (FFDM) imaging system. The films were then digitised using a flatbed scanner. The calibration curve relating the readings from a calibrated ionisation chamber and the films' net optical density (NOD) could not be obtained. The examination of the calibration data revealed non-sensitivity of the films to resolve dose differences below 20 mGy at 28 kVp. Therefore, radiochromic films were found not to be suitable for measuring ESD profiles in DM. A 2D map of the NOD of the irradiated films obtained using in-house developed MATLAB computer program is presented. PMID:25852183

  11. Comparison of dose accuracy between film and two-dimensional detectors in intensity-modulated radiation therapy

    NASA Astrophysics Data System (ADS)

    Onishi, Yuichi; Nakayama, Shinichi; Watanabe, Shinsaku; Kaneshige, Souichirou; Monzen, Hajime; Matsumoto, Kenji; Shintani, Naoya; Kamomae, Takeshi

    2015-07-01

    We constructed seven intensity-modulated radiation therapy (IMRT) treatment plans for prostate cancer (49 irradiation fields which contained seven randomly-sampled patients and seven fields) and evaluated the dose distributions by using a radiochromic film (EBT3 film) and a 2D detector. We superposed the calculated dose distribution of the IMRT treatment plan on EBT3 film and the 2D detector results and then compared those with the γ-analysis pass rate. The relative positions of the beam and the detector were varied; the results of the analysis of the superior-inferior (SI) direction potentially differed, depending on the detector position, under an irradiation beam with the same fluence map. The detector was moved over a range of' 8 mm in the SI direction in 1-mm step increments, measurement were made at each position, and the results were analyzed. The γ-analysis compared the dose distributions from EBT3 film and the radiation treatment planning system (RTPS) for each patient and field; the pass rate with the γ-analysis from 98 to 100% was 2.04%. When we compared the dose distributions of the 2D detector and the RTPS, the pass rate from 98 to 100% was 63.2%. The mean values for the ?-analysis pass rates for EBT3 film and the 2D detector were 94.2 and 97.6%, respectively. Volume averaging of the data indicated a mean pass rate and standard deviation of 98.6 and 0.91%, respectively, and a pass rate of more than 96% for all positions. A 2D detector can, therefore, be used as an alternative apparatus for IMRT dose verification.

  12. A feasibility study using radiochromic films for fast neutron 2D passive dosimetry

    PubMed Central

    Brady, Samuel L; Gunasingha, Rathnayaka; Yoshizumi, Terry T; Howell, Calvin R; Crowell, Alexander S; Fallin, Brent; Tonchev, Anton P; Dewhirst, Mark W

    2013-01-01

    The objective of this paper is threefold: (1) to establish sensitivity of XRQA and EBT radiochromic films to fast neutron exposure; (2) to develop a film response to radiation dose calibration curve and (3) to investigate a two-dimensional (2D) film dosimetry technique for use in establishing an experimental setup for a radiobiological irradiation of mice and to assess the dose to the mice in this setup. The films were exposed to a 10 MeV neutron beam via the 2H(d,n)3He reaction. The XRQA film response was a factor of 1.39 greater than EBT film response to the 10 MeV neutron beam when exposed to a neutron dose of 165 cGy. A film response-to-soft tissue dose calibration function was established over a range of 0–10 Gy and had a goodness of fit of 0.9926 with the calibration data. The 2D film dosimetry technique estimated the neutron dose to the mice by measuring the dose using a mouse phantom and by placing a piece of film on the exterior of the experimental mouse setup. The film results were benchmarked using Monte Carlo and aluminum (Al) foil activation measurements. The radiochromic film, Monte Carlo and Al foil dose measurements were strongly correlated, and the film within the mouse phantom agreed to better than 7% of the externally mounted films. These results demonstrated the potential application of radiochromic films for passive 2D neutron dosimetry. PMID:20693612

  13. NOTE: Fluorescent light effects on FWT-60 radiochromic film

    NASA Astrophysics Data System (ADS)

    Butson, Martin J.; Cheung, Tsang; Yu, Peter K. N.

    2005-08-01

    FWT-60 radiochromic film has been tested for colouration effects from fluorescent light sources and shown to produce a marked colouration when exposed to office fluorescent light sources showing an approximate 1 OD unit per 0.5 J m-2 exposure to a broad ultraviolet (UV) UVA + UVB spectrum at the peak absorption wavelength. This produces a measurable and quantifiable response to UV exposure. By choosing an appropriate wavelength of readout or band pass, the level of sensitivity can be changed to match the application or exposure level measurement required. These levels of UV response are significantly higher in sensitivity than other radiochromic films such as Gafchromic MD-55 by an order of magnitude. This feature may be of use for measurement of integrated UV exposure from fluorescent lights when required and produces a quantifiable history of total exposure.

  14. Stereotactic radiosurgery dosimetry using thermoluminescent dosimeters and radiochromic films

    NASA Astrophysics Data System (ADS)

    Ávila-Rodríguez, Miguel A.; Rodríguez-Villafuerte, Mercedes; Perches, Rodolfo Díaz

    2000-10-01

    In this work we present a protocol to measure absorbed dose distributions in stereotactic radiosurgery treatments with a linear accelerator (Linac) using thermoluminescent dosimeters (TLD) and radiochromic dye films. A Linac Philips SL-15 (6 MV X-rays), a ZD2 stereoactic system for localizing the target via computed tomography (CT), and Leibinger radiosurgery accessories will be used. A versatile spherical acrylic phantom of 16 cm diameter to measure the dose distributions has been designed. The phantom is composed of two hemispheres. On one flat side of the hemispheres an array of Harshaw/Bicron TLD-100 or a sheet of GafChromic MD-55 film will be placed. The phantom will be irradiated in three different orientations to obtain spatial dose distributions in the coronal, sagital and transverse planes. The experimental measurement will be compared with the results provided by a commercial treatment-planning system.

  15. A new correction method serving to eliminate the parabola effect of flatbed scanners used in radiochromic film dosimetry

    SciTech Connect

    Poppinga, D. Schoenfeld, A. A.; Poppe, B.; Doerner, K. J.; Blanck, O.; Harder, D.

    2014-02-15

    Purpose: The purpose of this study is the correction of the lateral scanner artifact, i.e., the effect that, on a large homogeneously exposed EBT3 film, a flatbed scanner measures different optical densities at different positions along thex axis, the axis parallel to the elongated light source. At constant dose, the measured optical densitiy profiles along this axis have a parabolic shape with significant dose dependent curvature. Therefore, the effect is shortly called the parabola effect. The objective of the algorithm developed in this study is to correct for the parabola effect. Any optical density measured at given position x is transformed into the equivalent optical density c at the apex of the parabola and then converted into the corresponding dose via the calibration of c versus dose. Methods: For the present study EBT3 films and an Epson 10000XL scanner including transparency unit were used for the analysis of the parabola effect. The films were irradiated with 6 MV photons from an Elekta Synergy accelerator in a RW3 slab phantom. In order to quantify the effect, ten film pieces with doses graded from 0 to 20.9 Gy were sequentially scanned at eight positions along thex axis and at six positions along the z axis (the movement direction of the light source) both for the portrait and landscape film orientations. In order to test the effectiveness of the new correction algorithm, the dose profiles of an open square field and an IMRT plan were measured by EBT3 films and compared with ionization chamber and ionization chamber array measurement. Results: The parabola effect has been numerically studied over the whole measuring field of the Epson 10000XL scanner for doses up to 20.9 Gy and for both film orientations. The presented algorithm transforms any optical density at positionx into the equivalent optical density that would be measured at the same dose at the apex of the parabola. This correction method has been validated up to doses of 5.2 Gy all over the

  16. Dose distribution around a needle-like anode X-ray tube: dye-film vs. planar thermoluminescent detectors.

    PubMed

    Budzanowski, M; Olko, P; Marczewska, B; Czopyk, L; Slapa, M; Stras, W; Traczyk, M; Talejko, M

    2006-01-01

    The dosimetry around the X-ray tube with a needle-like anode (NAXT), developed at the Institute of Nuclear Studies, for interstitial brachytherapy has been performed using (1) dye films (Gafchromic XR-T), (2) large-area thermoluminescent (TL) detectors--prepared either by gluing TL powder onto thin Al foil (so-called planar detectors with spatial resolution of 0.1 mm) and (3) miniature (2 mm diameter and 0.5 mm thick) TL detectors. The measurements were performed in following geometries. (1) Needle inside a PMMA cylinder--the planar TL detector mounted on the surface of the cylinder. (2) Needle inside a thick block of PMMA and TL detector mounted vertically 7 mm from needle axis. TL detectors were read with the planar (2D) thermoluminescence reader, developed at IFJ, with a sensitive CCD (charge couple device) camera. Gafchromic films were evaluated with a system based on Agfa Arcus 1200 scanner and calibrated with X rays (35 kV) filtered with 0.03 mm Mo and with Co-60 photons. The intensity distribution of TL light on the planar detector was calibrated in terms of absorbed dose to water, using (137)Cs gamma-rays. TL planar detectors seem to be a promising tool for 2D dosimetry of miniature X-ray sources. Obtained results for TLDs and Gafchromic films seem to be comparable but differences have been found. Both methods are useful for measurements of dose distribution around the NAXT X-rays source. PMID:16614087

  17. A comparison of TPS and different measurement techniques in small-field electron beams.

    PubMed

    Donmez Kesen, Nazmiye; Cakir, Aydin; Okutan, Murat; Bilge, Hatice

    2015-01-01

    In recent years, small-field electron beams have been used for the treatment of superficial lesions, which requires small circular fields. However, when using very small electron fields, some significant dosimetric problems may occur. In this study, dose distributions and outputs of circular fields with dimensions of 5cm and smaller, for nominal energies of 6, 9, and 15MeV from the Siemens ONCOR Linac, were measured and compared with data from a treatment planning system using the pencil-beam algorithm in electron beam calculations. All dose distribution measurements were performed using the Gafchromic EBT film; these measurements were compared with data that were obtained from the Computerized Medical Systems (CMS) XiO treatment planning system (TPS), using the gamma-index method in the PTW VeriSoft software program. Output measurements were performed using the Gafchromic EBT film, an Advanced Markus ion chamber, and thermoluminescent dosimetry (TLD). Although the pencil-beam algorithm is used to model electron beams in many clinics, there is no substantial amount of detailed information in the literature about its use. As the field size decreased, the point of maximum dose moved closer to the surface. Output factors were consistent; differences from the values obtained from the TPS were, at maximum, 42% for 6 and 15MeV and 32% for 9MeV. When the dose distributions from the TPS were compared with the measurements from the Gafchromic EBT films, it was observed that the results were consistent for 2-cm diameter and larger fields, but the outputs for fields of 1-cm diameter and smaller were not consistent. In CMS XiO TPS, calculated using the pencil-beam algorithm, the dose distributions of electron treatment fields that were created with circular cutout of a 1-cm diameter were not appropriate for patient treatment and the pencil-beam algorithm is not convenient for monitor unit (MU) calculations in electron dosimetry. PMID:25219322

  18. A comparison of TPS and different measurement techniques in small-field electron beams

    SciTech Connect

    Donmez Kesen, Nazmiye Cakir, Aydin; Okutan, Murat; Bilge, Hatice

    2015-04-01

    In recent years, small-field electron beams have been used for the treatment of superficial lesions, which requires small circular fields. However, when using very small electron fields, some significant dosimetric problems may occur. In this study, dose distributions and outputs of circular fields with dimensions of 5 cm and smaller, for nominal energies of 6, 9, and 15 MeV from the Siemens ONCOR Linac, were measured and compared with data from a treatment planning system using the pencil-beam algorithm in electron beam calculations. All dose distribution measurements were performed using the Gafchromic EBT film; these measurements were compared with data that were obtained from the Computerized Medical Systems (CMS) XiO treatment planning system (TPS), using the gamma-index method in the PTW VeriSoft software program. Output measurements were performed using the Gafchromic EBT film, an Advanced Markus ion chamber, and thermoluminescent dosimetry (TLD). Although the pencil-beam algorithm is used to model electron beams in many clinics, there is no substantial amount of detailed information in the literature about its use. As the field size decreased, the point of maximum dose moved closer to the surface. Output factors were consistent; differences from the values obtained from the TPS were, at maximum, 42% for 6 and 15 MeV and 32% for 9 MeV. When the dose distributions from the TPS were compared with the measurements from the Gafchromic EBT films, it was observed that the results were consistent for 2-cm diameter and larger fields, but the outputs for fields of 1-cm diameter and smaller were not consistent. In CMS XiO TPS, calculated using the pencil-beam algorithm, the dose distributions of electron treatment fields that were created with circular cutout of a 1-cm diameter were not appropriate for patient treatment and the pencil-beam algorithm is not convenient for monitor unit (MU) calculations in electron dosimetry.

  19. SU-E-CAMPUS-T-06: Radiochromic Film Analysis Based On Principal Components

    SciTech Connect

    Wendt, R

    2014-06-15

    Purpose: An algorithm to convert the color image of scanned EBT2 radiochromic film [Ashland, Covington KY] into a dose map was developed based upon a principal component analysis. The sensitive layer of the EBT2 film is colored so that the background streaks arising from variations in thickness and scanning imperfections may be distinguished by color from the dose in the exposed film. Methods: Doses of 0, 0.94, 1.9, 3.8, 7.8, 16, 32 and 64 Gy were delivered to radiochromic films by contact with a calibrated Sr-90/Y-90 source. They were digitized by a transparency scanner. Optical density images were calculated and analyzed by the method of principal components. The eigenimages of the 0.94 Gy film contained predominantly noise, predominantly background streaking, and background streaking plus the source, respectively, in order from the smallest to the largest eigenvalue. Weighting the second and third eigenimages by −0.574 and 0.819 respectively and summing them plus the constant 0.012 yielded a processed optical density image with negligible background streaking. This same weighted sum was transformed to the red, green and blue space of the scanned images and applied to all of the doses. The curve of processed density in the middle of the source versus applied dose was fit by a twophase association curve. A film was sandwiched between two polystyrene blocks and exposed edge-on to a different Y-90 source. This measurement was modeled with the GATE simulation toolkit [Version 6.2, OpenGATE Collaboration], and the on-axis depth-dose curves were compared. Results: The transformation defined using the principal component analysis of the 0.94 Gy film minimized streaking in the backgrounds of all of the films. The depth-dose curves from the film measurement and simulation are indistinguishable. Conclusion: This algorithm accurately converts EBT2 film images to dose images while reducing noise and minimizing background streaking. Supported by a sponsored research

  20. The performance of an optical cone-beam CT scanner adapted for radiochromic film dosimetry.

    PubMed

    Babic, Steven; Jordan, Kevin

    2012-11-01

    The primary purpose of this study was to evaluate commercial optical cone-beam computed tomography (CBCT) scanners as devices for reading EBT2 radiochromic film. A secondary objective was to implement a spatial correction for stray light present within optical CBCT systems. Square (12.7 × 12.7 cm²) EBT2 films were positioned vertically in the middle of a small water-filled tank, co-linear with the central beam axis of a 12 MeV electron beam. A total dose of 4.0 Gy was delivered at depth of 3.0 cm. Films were imaged prior to irradiation and 24 hours post-irradiation. Two different models of scanners, Vista15™ and Vista10™, were used to read out the irradiated films. In the Vista15™ scanner, residual light scatter was corrected for using: 1) a single vertical slot array and 2) a slot pair array that produced a vertical fan beam of light. Vista10™ was modified to have a smaller acceptance angle of scattered light and further corrections for residual scatter were made using a multiple slot array. With these different geometries, composite 'open field' and 'shadow field' images were generated and processed to create 'glare-free' pre and post-irradiation film images respectively, from which the net optical density (OD) was calculated. Results were compared against the open light field measurement in which no correction for stray light was made. Using the above scanners, EBT2 films were additionally read out to obtain 12 MeV electron and 6 MV photon percentage depth doses. By correcting for stray light it was found that the central-axis change in the net OD increased particularly in the 12 MeV electron build-up region and at the depth of maximum dose (d(max) = 3.0 cm) where light transmission is lowest. In the open light field measurement acquired with the Vista15™ scanner the net OD was 0.87 +/-0.02. Using single vertical slot array geometry to correct for stray light, the net OD was 0.94 +/-0.02, while with the slot pair array the net OD was 0.99 +/-0

  1. An investigation into factors affecting the precision of CT radiation dose profile width measurements using radiochromic films

    SciTech Connect

    Li, Baojun Behrman, Richard H.

    2015-04-15

    Purpose: To investigate the impact of x-ray beam energy, exposure intensity, and flat-bed scanner uniformity and spatial resolution on the precision of computed tomography (CT) beam width measurements using Gafchromic XR-QA2 film and an off-the-shelf document scanner. Methods: Small strips of Gafchromic film were placed at isocenter in a CT scanner and exposed at various x-ray beam energies (80–140 kVp), exposure levels (50–400 mA s), and nominal beam widths (1.25, 5, and 10 mm). The films were scanned in reflection mode on a Ricoh MP3501 flat-bed document scanner using several spatial resolution settings (100 to 400 dpi) and at different locations on the scanner bed. Reflection measurements were captured in digital image files and radiation dose profiles generated by converting the image pixel values to air kerma through film calibration. Beam widths were characterized by full width at half maximum (FWHM) and full width at tenth maximum (FWTM) of dose profiles. Dependences of these parameters on the above factors were quantified in percentage change from the baselines. Results: The uncertainties in both FWHM and FWTM caused by varying beam energy, exposure level, and scanner uniformity were all within 4.5% and 7.6%, respectively. Increasing scanner spatial resolution significantly increased the uncertainty in both FWHM and FWTM, with FWTM affected by almost 8 times more than FWHM (48.7% vs 6.5%). When uncalibrated dose profiles were used, FWHM and FWTM were over-estimated by 11.6% and 7.6%, respectively. Narrower beam width appeared more sensitive to the film calibration than the wider ones (R{sup 2} = 0.68 and 0.85 for FWHM and FWTM, respectively). The global and maximum local background variations of the document scanner were 1.2%. The intrinsic film nonuniformity for an unexposed film was 0.3%. Conclusions: Measurement of CT beam widths using Gafchromic XR-QA2 films is robust against x-ray energy, exposure level, and scanner uniformity. With proper film

  2. EBT ring physics

    SciTech Connect

    Uckan, N.A.

    1980-04-01

    This workshop attempted to evaluate the status of the current experimental and theoretical understanding of hot electron ring properties. The dominant physical processes that influence ring formation, scaling, and their optimal behavior are also studied. Separate abstracts were prepared for each of the 27 included papers. (MOW)

  3. Radiochromic films for dental CT dosimetry: a feasibility study.

    PubMed

    Rampado, O; Bianchi, S D; Peruzzo Cornetto, A; Rossetti, V; Ropolo, R

    2014-02-01

    Dental CT dose evaluations are commonly performed using thermoluminescent dosimeters (TLD) inside anthropomorphic phantoms. Radiochromic films with good sensitivity in the X-ray diagnostic field have recently been developed and are commercially available as GAFCHROMIC XR-QA. There are potential advantages in the use of radiochromic films such as a more comprehensive dosimetry thanks to the adjustable size of the film samples. The purpose of this study was to investigate the feasibility of using radiochromic films for dental CT dose evaluations. Film samples were cut with a width of 5mm and a length of 25 mm (strips), the same size as the Alderson Rando anthropomorphic phantom holes used in this study. Dental CT dose measurements were performed using simultaneously both TLD and radiochromic strips in the same phantom sites. Two equipment types were considered for dental CT examinations: a 16 slice CT and a cone beam CT. Organ equivalent doses were then obtained averaging the measurements from the sites of the same organ and effective doses were calculated using ICRP 103 weighting factors. The entire procedure was repeated four times for each CT in order to compare also the repeatability of the two dosimeter types. A linear correlation was found between the absorbed dose evaluated with radiochromic films and with TLD, with slopes of 0.930 and 0.944 (correlation r>0.99). The maximum difference between the two dosimeter's measurements was 25%, whereas the average difference was 7%. The measurement repeatability was comparable for the two dosimeters at cumulative doses above 15 mGy (estimated uncertainty at 1 sigma level of about 5%), whereas below this threshold radiochromic films show a greater dispersion of data, of about 10% at 1 sigma level. We obtained, using respectively Gafchromic and TLD measurements, effective dose values of 107 μSv and 117 μSv (i.e. difference of 8.6%) for the cone beam CT and of 523 μSv and 562 μSv (i.e. difference of 7%) for the

  4. Measurement of energy dependence for XRCT radiochromic film

    SciTech Connect

    Butson, Martin J.; Cheung, Tsang; Yu, Peter K. N.

    2006-08-15

    Gafchromic XRCT, radiochromic film is assessed over a broad energy range, from kilovoltage to megavoltage x rays for variations in reflected optical density to dose response. A large energy dependence was found with reflected optical density output for the same delivered dose varying from 7.8{+-}0.35 at 25.5 keV (50 kVp) peaking at 12.1{+-}0.5 at 54 keV (125 kVp) to 0.975{+-}0.03 at 2300 keV (10 MV) when normalized to 1 at 1400 keV (6 MV) energy. The response is constant (within 3%) in the 36-69 keV equivalent photon energy range, which corresponds to x-ray tube generating potentials of approximately 100-150 kVp. This matches well with beam qualities for diagnostic computed topography applications.

  5. SU-E-T-123: Dosimetric Comparison Between Portrait and Landscape Orientations in Radiochromic Film Dosimetry

    SciTech Connect

    Kakinohana, Y; Toita, T; Kasuya, G; Ariga, T; Heianna, J; Murayama, S

    2014-06-01

    Purpose: To compare the dosimetric properties of radiochromic films with different orientation. Methods: A sheet of EBT3 film was cut into eight pieces with the following sizes: 15×15 cm2 (one piece), 5x15 cm{sup 2} (two) and 4×5 cm{sup 2} (five). A set of two EBT3 sheets was used at each dose level. Two sets were used changing the delivered doses (1 and 2 Gy). The 5×15 cm{sup 2} pieces were rotated by 90 degrees in relation to each other, such that one had landscape orientation and the other had portrait orientation. All 5×15 cm2 pieces were irradiated with their long side aligned with the x-axis of the radiation field. The 15×15 cm{sup 2} pieces were irradiated rotated at 90 degrees to each other. Five pieces, (a total of ten from two sheets) were used to obtain a calibration curve. The irradiated films were scanned using an Epson ES-2200 scanner and were analyzed using ImageJ software. In this study, no correction was applied for the nonuniform scanner signal that is evident in the direction of the scanner lamp. Each film piece was scanned both in portrait and landscape orientations. Dosimetric comparisons of the beam profiles were made in terms of the film orientations (portrait and landscape) and scanner bed directions (perpendicular and parallel to the scanner movement). Results: In general, portrait orientation exhibited higher noise than landscape and was adversely affected to a great extent by the nonuniformity in the direction of the scanner lamp. A significant difference in the measured field widths between the perpendicular and parallel directions was found for both orientations. Conclusion: Without correction for the nonuniform scanner signal in the direction of the scanner lamp, a landscape orientation is preferable. A more detailed investigation is planned to evaluate quantitatively the effect of orientation on the dosimetric properties of a film.

  6. Dosimetry of dose distributions in radiotherapy of patients with surgical implants

    NASA Astrophysics Data System (ADS)

    Brożyna, Bogusław; Chełmiński, Krzysztof; Bulski, Wojciech; Giżyńska, Marta; Grochowska, Paulina; Walewska, Agnieszka; Zalewska, Marta; Kawecki, Andrzej; Krajewski, Romuald

    2014-11-01

    The investigation was performed in order to evaluate the use of Gafchromic EBT films for measurements of dose distributions created during radiotherapy in tissues surrounding titanium or resorbable implants used for joining and consolidating facial bones. Inhomogeneous dose distributions at implant-tissue interfaces can be the reason of normal tissue complications observed in radiotherapy patients after surgery with implants. The dose measured at a depth of 2.5 cm on contact surfaces, proximal and distal to the beam source, between the titanium implant and the phantom material was 109% and 92% respectively of the reference dose measured in a homogeneous phantom. For the resorbable implants the doses measured on the proximal and the distal contact surfaces were 102% and 101% respectively of the reference dose. The resorbable implants affect the homogeneity of dose distribution at a significantly lesser degree than the titanium implants. Gafchromic EBT films allowed for precise dose distribution measurements at the contact surfaces between tissue equivalent materials and implants. We measured doses at contact surfaces between titanium implants and RW3 phantom. We measured doses at contact surfaces between resorbable implants and RW3 phantom. We compared doses measured on contact surfaces and doses in homogeneous phantom. Doses at contact surfaces between RW3 phantom and titanium were distorted about 8-9%. Doses at RW3 phantom and resorbable implant contact surfaces were distorted about 2%.

  7. Dosimetry verification on VMAT and IMRT radiotherapy techniques: In the case of prostate cancer

    NASA Astrophysics Data System (ADS)

    Maulana, A.; Pawiro, S. A.

    2016-03-01

    Radiotherapy treatment depends on the accuracy of the dose delivery to patients, the purpose of the study is to verify the dose in IMRT and VMAT technique in prostate cancer cases correspond to TPS dose using phantom base on ICRU No.50. The dose verification of the target and OAR was performed by placing the TLD Rod LiF100 and EBT2 Gafchromic film at slab hole of pelvic part of the Alderson RANDO phantom for prostate cancer simulation. The Exposed TLDs was evaluated using the TLD Reader Harshaw while EBT2 film was scanned using Epson scanner. The point dose measurements were compared between planned dose and measured dose at target volume and OAR. The result is the dose difference at target volume, bladder and rectum for IMRT and VMAT are less than 5%. On the other hand, the dose difference at the Femoral head is more than 5% for both techniques because the location of OAR already in low gradient dose. Furthermore, the difference dose of the target volume for IMRT technique tends to be smaller than VMAT either for TLD and EBT2 film detectors. From the measurement showed that the delivered dose on the phantom simulation match with ICRU No.50 criteria.

  8. Film Reviews.

    ERIC Educational Resources Information Center

    Lance, Larry M.; Atwater, Lynn

    1987-01-01

    Reviews four Human Sexuality films and videos. These are: "Personal Decisions" (Planned Parenthood Federation of America, 1985); "The Touch Film" (Sterling Production, 1986); "Rethinking Rape" (Film Distribution Center, 1985); "Not A Love Story" (National Film Board of Canada, 1981). (AEM)

  9. Optically stimulated luminescence (OSL) of carbon-doped aluminum oxide (Al{sub 2}O{sub 3}:C) for film dosimetry in radiotherapy

    SciTech Connect

    Schembri, V.; Heijmen, B. J. M.

    2007-06-15

    Introduction and Purpose: Conventional x-ray films and radiochromic films have inherent challenges for high precision radiotherapy dosimetry. Here we have investigated basic characteristics of optically stimulated luminescence (OSL) of irradiated films containing carbon-doped aluminum oxide (Al{sub 2}O{sub 3}:C) for dosimetry in therapeutic photon and electron beams. Materials and Methods: The OSL films consist of a polystyrene sheet, with a top layer of a mixture of single crystals of Al{sub 2}O{sub 3}:C, ground into a powder, and a polyester base. The total thickness of the films is 0.3 mm. Measurements have been performed in a water equivalent phantom, using 4, 6, 10, and 18 MV photon beams, and 6-22 MeV electron beams. The studies include assessment of the film response (acquired OSL signal/delivered dose) on delivered dose (linearity), dose rate (1-6 Gy/min), beam quality, field size and depth (6 MV, ranges 4x4-30x30 cm{sup 2}, d{sub max}-35 cm). Doses have been derived from ionization chamber measurements. OSL films have also been compared with conventional x-ray and GafChromic films for dosimetry outside the high dose area, with a high proportion of low dose scattered photons. In total, 787 OSL films have been irradiated. Results: Overall, the OSL response for electron beams was 3.6% lower than for photon beams. Differences between the various electron beam energies were not significant. The 6 and 18 MV photon beams differed in response by 4%. No response dependencies on dose rate were observed. For the 6 MV beam, the field size and depth dependencies of the OSL response were within {+-}2.5%. The observed inter-film response variation for films irradiated with the same dose varied from 1% to 3.2% (1 SD), depending on the measurement day. At a depth of 20 cm, 5 cm outside the 20x20 cm{sup 2} 6 and 18 MV beams, an over response of 17% was observed. In contrast to GafChromic and conventional x-ray films, the response of the Al{sub 2}O{sub 3}:C films is linear

  10. Radiochromic film sensitivity calibrations using ion beams from a Pelletron accelerator

    NASA Astrophysics Data System (ADS)

    Filkins, T. M.; Steidle, Jessica; Ward, R. J.; Freeman, C. G.; Padalino, S. J.; Regan, S. P.; Sangster, T. C.

    2015-11-01

    Radiochromic film (RCF) is a transparent detector film that permanently changes color following exposure to ionizing radiation. The optical density of the film increases with increasing absorbed dose. RCF is convenient to use because it requires no chemical processing and can be scanned using commercially available document scanners. RCF is used frequently in medical applications, but is also used in a variety of diagnostics in high energy density physics. The film consists of a single or double layer of radiation-sensitive organic microcrystal monomers placed onto a polyester backing. GafchromicTM manufactures a large number of different types of RCF, and new types of film frequently replace older products. In this study, the sensitivity of several types of RCF to ion beams of different energies was measured. Ion beams produced by the SUNY Geneseo 1.7 MV Pelletron accelerator were directed into a target chamber where they scattered off of a gold foil. A sample of RCF was exposed to the scattered ions. The fluence of incident particles on the film was measured using a surface barrier detector. Results of these calibrations will be presented. This work was funded in part by a grant from the DOE through the Laboratory for Laser Energetics.

  11. Small Radiation Beam Dosimetry for Radiosurgery of Trigeminal Neuralgia: One Case Analysis

    SciTech Connect

    Garcia-Garduno, O. A.; Larraga-Gutierrez, J. M.; Rodriguez-Villafuerte, M.; Martinez-Davalos, A.; Moreno-Jimenez, S.; Suarez-Campos, J. J.; Celis, M. A.

    2008-08-11

    The use of small radiation beams for trigeminal neuralgia (TN) treatment requires high precision and accuracy in dose distribution calculations and delivery. Special attention must be kept on the type of detector to be used. In this work, the use of GafChromic EBT registered radiochromic and X-OMAT V2 radiographic films for small radiation beam characterization is reported. The dosimetric information provided by the films (total output factors, tissue maximum ratios and off axis ratios) is compared against measurements with a shielded solid state (diode) reference detector. The film dosimetry was used for dose distribution calculations for the treatment of trigeminal neuralgia radiosurgery. Comparison of the isodose curves shows that the dosimetry produced with the X-OMAT radiographic film overestimates the dose distributions in the penumbra region.

  12. Solid-state fermentation: tool for bioremediation of adsorbed textile dyestuff on distillery industry waste-yeast biomass using isolated Bacillus cereus strain EBT1.

    PubMed

    Kadam, Avinash A; Kamatkar, Jeevan D; Khandare, Rahul V; Jadhav, Jyoti P; Govindwar, Sanjay P

    2013-02-01

    Bioremediation of textile dyestuffs under solid-state fermentation (SSF) using industrial wastes as substrate pose an economically feasible, promising, and eco-friendly alternative. The purpose of this study was to adsorb Red M5B dye, a sample of dyes mixture and a real textile effluent on distillery industry waste-yeast biomass (DIW-YB) and its further bioremediation using Bacillus cereus EBT1 under SSF. Textile dyestuffs were allowed to adsorb on DIW-YB. DIW-YB adsorbed dyestuffs were decolorized under SSF by using B. cereus. Enzyme analysis was carried out to ensure decolorization of Red M5B. Metabolites after dye degradation were analyzed using UV-Vis spectroscopy, FTIR, HPLC, and GC-MS. DIW-YB showed adsorption of Red M5B, dyes mixture and a textile wastewater sample up to 87, 70, and 81 %, respectively. DIW-YB adsorbed Red M5B was decolorized up to 98 % by B. cereus in 36 h. Whereas B. cereus could effectively reduce American Dye Manufacture Institute value from DIW-YB adsorbed mixture of textile dyes and textile wastewater up to 70 and 100 %, respectively. Induction of extracellular enzymes such as laccase and azoreductase suggests their involvement in dye degradation. Repeated utilization of DIW-YB showed consistent adsorption and ADMI removal from textile wastewater up to seven cycles. HPLC and FTIR analysis confirms the biodegradation of Red M5B. GC-MS analysis revealed the formation of new metabolites. B. cereus has potential to bioremediate adsorbed textile dyestuffs on DIW-YB. B. cereus along with DIW-YB showed enhanced decolorization performance in tray bioreactor which suggests its potential for large-scale treatment procedures. PMID:22562346

  13. Dosimetry of Strontium eye applicator: Comparison of Monte Carlo calculations and radiochromic film measurements

    NASA Astrophysics Data System (ADS)

    Laoues, M.; Khelifi, R.; Moussa, A. S.

    2015-01-01

    Strontium-90 eye applicators are a beta-ray emitter with a relatively high-energy (maximum energy about 2.28 MeV and average energy about 0.9 MeV). These applicators come in different shapes and dimensions; they are used for the treatment of eye diseases. Whenever, radiation is used in treatment, dosimetry is essential. However, knowledge of the exact dose distribution is a critical decision-making to the outcome of the treatment. The main aim of our study is to simulate the dosimetry of the SIA.20 eye applicator with Monte Carlo GATE 6.1 platform and to compare the calculated results with those measured with EBT2 films. This means that GATE and EBT2 were used to quantify the surface and depths dose- rate, the relative dose profile and the dosimetric parameters in according to international recommendations. Calculated and measured results are in good agreement and they are consistent with the ICRU and NCS recommendations.

  14. Poster — Thur Eve — 20: CTDI Measurements using a Radiochromic Film-based clinical protocol

    SciTech Connect

    Quintero, C.; Bekerat, H.; DeBlois, F.; Tomic, N.; Devic, S.; Seuntjens, J.

    2014-08-15

    The purpose of the study was evaluating accuracy and reproducibility of a radiochromic film-based protocol to measure computer tomography dose index (CTDI) as a part of annual QA on CT scanners and kV-CBCT systems attached to linear accelerators. Energy dependence of Gafchromic XR-QA2 ® film model was tested over imaging beam qualities (50 – 140 kVp). Film pieces were irradiated in air to known values of air-kerma (up to 10 cGy). Calibration curves for each beam quality were created (Film reflectance change Vs. Air-kerma in air). Film responses for same air-kerma values were compared. Film strips were placed into holes of a CTDI phantom and irradiated for several clinical scanning protocols. Film reflectance change was converted into dose to water and used to calculate CTDIvol values. Measured and tabulated CTDIvol values were compared. Average variations of ±5.2% in the mean film reflectance change were observed in the energy range of 80 to 140 keV, and 11.1% between 50 and 140 keV. Measured CTDI values were in average 10% lower than tabulated CTDI values for CT-simulators, and 44% higher for CBCT systems. Results presented a mean variation for the same machine and protocol of 2.6%. Variation of film response is within ±5% resulting in ±15% systematic error in dose estimation if a single calibration curve is used. Relatively large discrepancy between measured and tabulated CTDI values strongly support the trend towards replacing CTDI value with equilibrium dose measurement in the center of cylindrical phantom, as suggested by TG- 111.

  15. Dosimetry of biological irradiations using radiochromic films

    NASA Astrophysics Data System (ADS)

    Aldelaijan, S.; Nobah, A.; Alsbeih, G.; Moftah, B.; Aldahlawi, I.; Alzahrany, A.; Tomic, N.; Devic, S.

    2013-05-01

    Delivering accurate radiation dose to blood specimens during biological irradiations is essential in quantifying damage of ionizing radiation. To estimate dose to blood samples as accurately as possible, pieces of EBT2 model GAFCHROMIC™ film were placed within an approximately 10 mm finely ground rice layer that was used to simulate test specimens inside 40 mL plastic flasks. Irradiations of flasks were carried out using an X-RAD 320 irradiator with a beam quality of 320 kVp and a measured half value layer of 1.12 mm Cu, in air and in a full scattering setup which consisted of either rice or Solid Water™ (SW) surrounding flasks, filled to the same level at top of the flasks, together with a 5 cm thick SW slab beneath them. Outputs, per cent depth doses and beam profiles at different depths were measured and compared between setups. For the same setting, the dose delivered to the middle flask under the full scattering setup is 22% larger than with the in-air setup at the depth of the specimen and 9.2% more homogeneous across the specimen thickness of 10 mm (2.3% variation in comparison to the surface). Rice showed a fairly similar performance to SW within 1% at the same depth of 10 mm. Experimental setup based on full scattering conditions was shown to provide faster, more homogenous and fairly uniform dose delivery to biological specimens in comparison to conventionally used in-air setups.

  16. TH-C-18A-05: Characteristics and Accuracy of CT Radiation Profile Width Measurements Using Radiochromic Films

    SciTech Connect

    Li, B; Behrman, R

    2014-06-15

    Purpose: To investigate the impact of several acquisition and image analysis factors on the accuracy of CT beam width measurement using Gafchromic XR-QA2 film and an off the shelf document scanner. Methods: The film was carefully placed at the iso-center and exposed at various x-ray beam energies (80–140 kVp), exposure levels (50–400 mAs), and nominal beam widths (1.25, 5, and 10 mm). Films were scanned on a RICOH MP3501 flat-bed document scanner using reflection mode and various spatial resolutions (100 to 600 dpi) and locations within the scanner bed. Radiation dose profiles were derived by converting the net reflection measurements to dose values (in mGy) through film calibration (which is also described here). The beam widths were characterized by the FWHM and FWTM of the dose profiles and the deviations from the hybrid reference standards were quantified. The impact of x-ray beam energy, exposure intensity, film calibration, and the uniformity and spatial resolution of the document scanner were investigated. Results: The uncertainties in both FWHM and FWTM caused by varying beam energy, exposure level, and scanner uniformity were all within approximately 5%. Increasing scanner spatial resolution increased the uncertainty in both FWHM and FWTM (R2 = 0.35–0.71), with the FWTM affected by more than five times than the FWHM (28% vs. 4.9%). When uncalibrated dose profiles were used, the FWHM and FWTM were over-estimated by as much as 11.6% and 7.6%, respectively. Narrower beam width was more sensitive to the film calibration than the wider one (R2 = 0.68 and 0.85 for the FWHM and FWTM, respectively). Conclusion: Measurement of CT beam width using Gafchromic XR-QA2 films is robust against x-ray energy, exposure level, and scanner uniformity. It can provide adequate accuracy for all practical purposes with appropriate film calibration and scanner spatial resolution (100 dpi or less)

  17. Is It Possible to Publish a Calibration Function for Radiochromic Film?

    PubMed Central

    Chan, Maria F.; Lewis, David; Yu, Xiang

    2016-01-01

    Purpose To assess the possibility of using a public calibration function for radiochromic film dosimetry in dose QA of highly conformal treatment plans. Methods EBT3 film calibration strips (3.5 × 20 cm2 from lots A101212 and A011713) were exposed on a Varian Trilogy at a facility to a 10 × 10 cm2 open field at doses of 80, 160, 320 cGy using 6MV photons. Together with a strip of unexposed film from the same lot the exposed films were digitized in a single scan using different Epson 10,000 XL scanners at two different facilities. The dose-response data for each color-channel from each facility were generated using the same calibration function X(D) = a + b/(D − c), where X(D) is the response at dose D and a, b and c are the coefficients. Different batches of EBT3 film were exposed to a VMAT beam. These films, plus two reference strips exposed to doses of zero and 160 cGy, were digitized on the scanners at the two facilities. Using the multi-channel dosimetry method and One-scan protocol (Med Phys, 39:6339–49, 2012) the recorded doses on the VMAT films were calculated and the results were compared with the VMAT plan using a Gamma index of 3%/3 mm. Results The passing rates obtained for dose maps calculated for all combinations of VMAT images and calibration functions were nearly unchanged, using the One-scan protocol. Also, in all cases a passing rate of >99% was obtained for Gamma index of 3%/3 mm. On the other hand, if the One-scan protocol was not employed, the dose maps for VMAT images and calibration functions from different scanners showed poor correlation with the treatment plan. This is probably due to the scan-to-scan variability. Conclusions The authors have found that it is feasible to use a public calibration function for a given radiochromic film lot using the same methodology, One-scan protocol, for patient-specific QA.

  18. Measuring energy response for RTQA radiochromic film to improve quality assurance procedures.

    PubMed

    Butson, M J; Cheung, T; Yu, P K N

    2008-09-01

    RTQA Gafchromic, radiochromic film is assessed for its radiation energy dependence in photon beams ranging from superficial to megavoltage energies. RTQA radiochromic film has uses in radiation quality assurance procedures due to its auto development and visualisation properties. These properties allow for immediate comparison of x-ray alignment and coincidence not available with radiographic films. Results show that the RTQA film produces an energy dependant darkening to x-rays which results in x-ray energies of 69 keV photon equivalent (150 kVp) to produce 2.14 times the optical density to dose ratio of a 6MV x-ray beam. The following dose ratio's (normalized to 1 at 150 kVp) provide the same net optical density change for RTQA film. 1.47-50 kVp : 1.21-75 kVp : 1.09-100 kVp : 1.01-125 kVp: 1.00-150 kVp : 1.03-200 kVp : 1.07-250 kVp : 2.14-6 MVp : 2.14 10 MVp. Although the film is not designed to be used as a quantitative measure of radiation it is still useful to know its energy response at differing x-ray energies to expose the film to the appropriate dose to provide optimal darkening characteristics for a given QA test at the appropriate energy. Our results have shown that a 0.3 optical density change with RTQA film provides a colour change level useable for accurate alignment procedures PMID:18946978

  19. MCNP modelling of vaginal and uterine applicators used in intracavitary brachytherapy and comparison with radiochromic film measurements

    NASA Astrophysics Data System (ADS)

    Ceccolini, E.; Gerardy, I.; Ródenas, J.; van Dycke, M.; Gallardo, S.; Mostacci, D.

    Brachytherapy is an advanced cancer treatment that is minimally invasive, minimising radiation exposure to the surrounding healthy tissues. Microselectron© Nucletron devices with 192Ir source can be used for gynaecological brachytherapy, in patients with vaginal or uterine cancer. Measurements of isodose curves have been performed in a PMMA phantom and compared with Monte Carlo calculations and TPS (Plato software of Nucletron BPS 14.2) evaluation. The isodose measurements have been performed with radiochromic films (Gafchromic EBT©). The dose matrix has been obtained after digitalisation and use of a dose calibration curve obtained with a 6 MV photon beam provided by a medical linear accelerator. A comparison between the calculated and the measured matrix has been performed. The calculated dose matrix is obtained with a simulation using the MCNP5 Monte Carlo code (F4MESH tally).

  20. SU-E-T-223: Investigation of the Accuracy of Two-Dimensional Dose Distributions Measurement From High-Dose-Rate Brachytherapy Ir-192 Source Using Multiple-Diode-Array Detector (MapCheck2)

    SciTech Connect

    Taguenang, J; De La Fuente, T Herman; Ahmad, S; Ali, I

    2014-06-01

    Purpose: To investigate the dosimetric accuracy of multiple-diode-array detector (Mapcheck2) for high-dose-rate brachytherapy Ir-192 source. The two-dimensional (2D) dose distributions measured with MapCheck2 were validated with EBT2 Gafchromic film measurement and AAPM task-group- 43 (TG-43) modeling. Methods: 2D-dose distributions from Ir-192 source were measured with MapCheck2 and EBT2-films. MapCheck2 response was corrected for effects: directional dependence, diode and phantom heterogeneity. Optical density growth of the film was controlled by synchronized scanning of the film exposed to Ir-192 and calibration films exposed to 6 MV linac beams. Similarly, MapCheck2 response was calibrated to dose using 6 MV beams. An empirical model was developed for the dose distributions measured with Mapcheck2 that considered directional, diode and phantom heterogeneity corrections. The dose deposited in solid-state-detectors was modeled using a cavity theory model for the diode. This model was then validated with measurements using EBT2-films and calculations with TG-43. Results: The response of MapCheck2 has been corrected for different effects including: (a) directional dependence of 0–20% over angular range 0o–90o, (b) phantom heterogeneity (3%) and (c) diode heterogeneity (9%). The corrected dose distributions measured with MapCheck2 agreed well with the measured dose distributions from EBT2-film and with calculations using TG-43 within 5% over a wide range of dose levels and rates. The advantages of MapCheck2 include less noisy, linear and stable response compared with film. The response of MapCheck2 exposed to 192Ir-source showed no energy dependence similar to its response to MV energy beam. Detection spatial-resolution of individual diodes was 0.8×0.8 mm2, however, 2DMapCheck2 resolution is limited by distance between diodes (7.07 mm). Conclusion: The dose distribution measured with MapCheck2 agreed well within 5% with that measured using EBT2-films; and

  1. Unconventional heating experiments in EBT

    SciTech Connect

    Glowienka, J.C.; Davis, W.A.; Hillis, D.L.; Uckan, T.; Bieniosek, F.M.; Solensten, L.

    1982-04-01

    Reported here are the results of two nonstandard heating experiments. In one, the fundamental ECH resonance at either 10.6 or 18 GHz was placed at the usual ring position in an attempt to generate rings solely with fundamental heating; this is the fundamental resonance configuration (FR). In a second experiment, an attempt was made to heat the ring and core separately by forming the ring as above with 10.6 GHz and then heating the core with the fundamental and second harmonic of 28 GHz; this is the separately coupled ring and toroidal core heating configuration (SCRATCH). In both cases, the hot electron annuli were formed and stabilized the core plasma as in the SR experiments. In FR, the rings form at the second harmonic and no significant potential structures were found. In SCRATCH, either a potential well or a potential hill could be formed. In both heating regimes, significant ion heating above that found in standard heating experiments was found.

  2. Micrometer-resolved film dosimetry using a microscope in microbeam radiation therapy

    SciTech Connect

    Bartzsch, Stefan Oelfke, Uwe; Lott, Johanna; Welsch, Katrin; Bräuer-Krisch, Elke

    2015-07-15

    Purpose: Microbeam radiation therapy (MRT) is a still preclinical tumor therapy approach that uses arrays of a few tens of micrometer wide parallel beams separated by a few 100 μm. The production, measurement, and planning of such radiation fields are a challenge up to now. Here, the authors investigate the feasibility of radiochromic film dosimetry in combination with a microscopic readout as a tool to validate peak and valley doses in MRT, which is an important requirement for a future clinical application of the therapy. Methods: Gafchromic{sup ®} HD-810 and HD-V2 films are exposed to MRT fields at the biomedical beamline ID17 of the European Synchrotron Radiation Facility (ESRF) and are afterward scanned with a microscope. The measured dose is compared with Monte Carlo calculations. Image analysis tools and film handling protocols are developed that allow accurate and reproducible dosimetry. The performance of HD-810 and HD-V2 films is compared and a detailed analysis of the resolution, noise, and energy dependence is carried out. Measurement uncertainties are identified and analyzed. Results: The dose was measured with a resolution of 5 × 1000 μm{sup 2} and an accuracy of 5% in the peak and between 10% and 15% in the valley region. As main causes for dosimetry uncertainties, statistical noise, film inhomogeneities, and calibration errors were identified. Calibration errors strongly increase at low doses and exceeded 3% for doses below 50 and 70 Gy for HD-V2 and HD-810 films, respectively. While the grain size of both film types is approximately 2 μm, the statistical noise in HD-V2 is much higher than in HD-810 films. However, HD-810 films show a higher energy dependence at low photon energies. Conclusions: Both film types are appropriate for dosimetry in MRT and the microscope is superior to the microdensitometer used before at the ESRF with respect to resolution and reproducibility. However, a very careful analysis of the image data is required

  3. Film Boxes.

    ERIC Educational Resources Information Center

    Osterer, Irv

    2002-01-01

    Presents an art lesson in which students created three-dimensional designs for 35mm film packages to improve graphic arts learning. Describes how the students examined and created film boxes using QuarkXPress software. (CMK)

  4. Film Reviews

    ERIC Educational Resources Information Center

    Dowling, John, Ed.

    1976-01-01

    Reviews five instructional films on: P-N junctions; crystal diodes; nuclear fusion research; Schlieren photography; and the energy crisis; including discussions of solar, nuclear, and fossil fuel energy. Also lists numerous other available films. (MLH)

  5. SU-F-BRE-03: Consideration of a Track-Interaction Model for Radiochromic Film Response

    SciTech Connect

    Rosen, B; DeWerd, L

    2014-06-15

    Purpose: Conventional methods for characterizing the energy response of radiochromic film (RCF) typically involve assessing changes in response when exposed to various beam qualities and use Monte Carlo to determine absorbed dose. These methods represent RCF as a dose integrator of a homogeneous energy deposition volume. Apparent film saturation, nonlinearity, and intrinsic energy dependence are unpredicted with conventional methods. Recent work has shown significant RCF intrinsic energy dependence, which limits its use in absolute dosimetry. This work introduces a track-interaction model (TIM) for RCF and assesses its ability to predict total energy response. Methods: A TIM based on Katz single-hit theory was developed to accumulate energy flux along particle tracks within active crystals, represented as (1×1×20)um{sup 3} prisms about the Gafchromic™ EBT3 active volume using MCNP5 and Matlab. Energy flux contributed to film response only if near the threshold energy for polymerization in polydiacetylenes (2.5eV/monomer). Energy deposition in excess of maximum efficiency represented crystal saturation and did not contribute to film response. The TIM was applied to RCF exposed in air to various monoenergetic photon beams and Co-60. Geometric distribution of energy flux was found for each beam quality in a (1×1)mm{sup 2} RCF area. RCF response relative to Co-60 absorbed dose-to-water (S-TIM) was determined and compared to published values (S-PUB). Results: TIM successfully predicted that lower energy radiation is less effective at inducing polymerization, though the magnitude of the phenomenon was overpredicted. S-TIM was −29% and +20% for 20 and 40 keV, respectively. This agreed qualitatively with S-PUB of −27% and +16%. TIM-generated sensitometric curves contained the non-linearity and saturation apparent in RCF. Conclusion: This work indicates the possibility for TIMs to predict changes in RCF response to various energies. Future work will refine TIM by

  6. Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry

    SciTech Connect

    Lewis, David; Chan, Maria F.

    2015-01-15

    Purpose: A known factor affecting the accuracy of radiochromic film dosimetry is the lateral response artifact (LRA) induced by nonuniform response of a flatbed scanner in the direction perpendicular to the scan direction. This work reports a practical solution to eliminate such artifacts for all forms of dose QA. Methods: EBT3 films from a single production lot (02181401) cut into rectangular 4 × 5 cm{sup 2} pieces, with the long dimension parallel to the long dimension of the original 20.3 × 25.4 cm{sup 2} sheets, were exposed at a depth of 5 cm on a Varian Trilogy at the center of a 20 × 20 cm{sup 2} open field at seven doses between 50 and 1600 cGy using 6 MV photons. These films together with an unexposed film from the same production lot were lined one next to the other on an Epson 10000XL or 11000XL scanner in portrait orientation with their long dimension parallel to the scan direction. Scanned images were then obtained with the line of films positioned at seven discrete lateral locations perpendicular to the scan direction. The process was repeated in landscape orientation and on three other Epson scanners. Data were also collected for three additional production lots of EBT3 film (11051302, 03031401, and 03171403). From measurements at the various lateral positions, the scanner response was determined as a function of the lateral position of the scanned film. For a given color channel X, the response at any lateral position L is related to the response at the center, C, of the scanner by Response(C, D, X) = A{sub L,X} + B{sub L,X}⋅Response(L, D, X), where D is dose and the coefficients A{sub L,X} and B{sub L,X} are determined from the film measurements at the center of the scanner and six other discrete lateral positions. The values at intermediate lateral positions were obtained by linear interpolation. The coefficients were determined for the red, green, and blue color channels, preserving the ability to apply triple-channel dosimetry once

  7. Nanocomposite films

    DOEpatents

    Mitlin, David; , Ophus, Colin; Evoy, Stephane; Radmilovic, Velimir; Mohammadi, Reza; Westra, Ken; Nelson-Fitzpatrick, Nathaniel; Lee, Zonghoon

    2010-07-20

    A thin-film composition of nanocrystal molybdenum in an amorphous metallic matrix may be formed by co-sputtering Mo with aluminum or nickel. NEMS cantilevers may be formed from the film. The films exhibit high nanoindentation hardness and a reduction in roughness and intrinsic stress, while maintaining resistivity in the metallic range.

  8. On Film

    ERIC Educational Resources Information Center

    Watts, Marty

    2006-01-01

    In this article, the author discusses the role of window films in enhancing indoor air quality in schools. Historically, window film has been used to reduce temperatures in buildings prone to overheating. Too much solar energy entering through windows makes occupants uncomfortable and air conditioning more costly. Film has been a simple solution…

  9. Humanistic Films.

    ERIC Educational Resources Information Center

    Gaffney, Maureen, Ed.

    1981-01-01

    Designed for media specialists and educators, this issue includes seven articles focusing on humanistic films for children. Following a brief editorial encouraging the ideals of humanism, the first article presents an analysis of seven films with positive sex-role models. Included is a model for evaluating children's films. The second article…

  10. Build-up material requirements in clinical dosimetry during total body irradiation treatments

    PubMed Central

    Butson, Martin; Pope, Dane; Haque, Mamoon; Chen, Tom; Song, Guangli; Whitaker, May

    2016-01-01

    Total body irradiation (TBI) treatments are mainly used in a preparative regimen for hematopoietic stem cell (or bone marrow) transplantation. Our standard clinical regimen is a 12 Gy/6 fraction bi-daily technique using 6MV X-rays at a large extended source to surface distance (SSD). This work investigates and quantifies the dose build-up characteristics and thus the requirements for bolus used for in vivo dosimetry for TBI applications. Percentage dose build-up characteristics of photon beams have been investigated at large extended SSDs using ionization chambers and Gafchromic film. Open field measurements at different field sizes and with differing scatter conditions such as the introduction of standard Perspex scattering plates at different distances to the measurement point were made in an effort to determine the required bolus/build-up material required for accurate determination of applied dose. Percentage surface dose values measured for open fields at 300 cm SSD were found to range from 20% up to 65.5% for fields 5 cm × 5 cm to 40 cm × 40 cm, respectively. With the introduction of 1 cm Perspex scattering plates used in TBI treatments, the surface dose values increased up to 83–90% (93–97% at 1 mm depth), depending on the position of the Perspex scattering plate compared to the measurement point. Our work showed that at least 5 mm water equivalent bolus/scatter material should be placed over the EBT3 film for accurate dose assessment for TBI treatments. Results also show that a small but measurable decrease in measured dose occurred with 5 mm water equivalent thick bolus material of areas '3 cm2. As such, we recommend that 3 cm × 3 cm × 5 mm bolus build-up is the smallest size that should be placed over EBT3 Gafchromic film when used for accurate in vivo dosimetry for TBI applications. PMID:27217628

  11. Build-up material requirements in clinical dosimetry during total body irradiation treatments.

    PubMed

    Butson, Martin; Pope, Dane; Haque, Mamoon; Chen, Tom; Song, Guangli; Whitaker, May

    2016-01-01

    Total body irradiation (TBI) treatments are mainly used in a preparative regimen for hematopoietic stem cell (or bone marrow) transplantation. Our standard clinical regimen is a 12 Gy/6 fraction bi-daily technique using 6MV X-rays at a large extended source to surface distance (SSD). This work investigates and quantifies the dose build-up characteristics and thus the requirements for bolus used for in vivo dosimetry for TBI applications. Percentage dose build-up characteristics of photon beams have been investigated at large extended SSDs using ionization chambers and Gafchromic film. Open field measurements at different field sizes and with differing scatter conditions such as the introduction of standard Perspex scattering plates at different distances to the measurement point were made in an effort to determine the required bolus/build-up material required for accurate determination of applied dose. Percentage surface dose values measured for open fields at 300 cm SSD were found to range from 20% up to 65.5% for fields 5 cm × 5 cm to 40 cm × 40 cm, respectively. With the introduction of 1 cm Perspex scattering plates used in TBI treatments, the surface dose values increased up to 83-90% (93-97% at 1 mm depth), depending on the position of the Perspex scattering plate compared to the measurement point. Our work showed that at least 5 mm water equivalent bolus/scatter material should be placed over the EBT3 film for accurate dose assessment for TBI treatments. Results also show that a small but measurable decrease in measured dose occurred with 5 mm water equivalent thick bolus material of areas '3 cm(2). As such, we recommend that 3 cm × 3 cm × 5 mm bolus build-up is the smallest size that should be placed over EBT3 Gafchromic film when used for accurate in vivo dosimetry for TBI applications. PMID:27217628

  12. Evaluation of spatial resolution in image acquisition by optical flatbed scanners for radiochromic film dosimetry

    NASA Astrophysics Data System (ADS)

    Asero, G.; Greco, C.; Gueli, A. M.; Raffaele, L.; Spampinato, S.

    2016-03-01

    Introduction: Radiochromic films are two-dimensional dosimeters that do not require developing and give values of absorbed dose with accuracy and precision. Since this dosimeter colours directly after irradiation, it can be digitized with commercial optical flatbed scanners to obtain a calibration curve that links blackening of the film with dose. Although the film has an intrinsic high spatial resolution, the scanner determines the actual resolution of this dosimeter, in particular the "dot per inch" (dpi) parameter. The present study investigates the effective spatial resolution of a scanner used for Gafchromic® XR-QA2 film (designed for radiology Quality Assurance) analysis. Material and methods: The quantitative evaluation of the resolution was performed with the Modulation Transfer Function (MTF) method, comparing the nominal resolution with the experimental one. The analysis was performed with two procedures. First, the 1951 USAF resolution test chart, a tool that tests the performance of optical devices, was used. Secondly, a combined system of mammography X-ray tube, XR-QA2 film and a bar pattern object was used. In both cases the MTF method has been applied and the results were compared. Results: The USAF and the film images have been acquired with increasing dpi and a standard protocol for radiochromic analysis, to evaluate horizontal and vertical and resolution. The effective resolution corresponds to the value of the MTF at 50%. In both cases and for both procedures, it was verified that, starting from a dpi value, the effective resolution saturates. Conclusion: The study found that, for dosimetric applications, the dpi of the scanner have to be adjusted to a reasonable value because, if too high, it requires high scanning and computational time without providing additional information.

  13. Comparison of small photon beams measured using radiochromic and silver-halide films in solid water phantoms

    SciTech Connect

    Zeidan, Omar A.; Li, Jonathan G.; Low, Daniel A.; Dempsey, James F.

    2004-10-01

    In this study, we compared the dosimetric properties of four of the most commonly used films for megavoltage photon-beam dosimetry when irradiated under identical conditions by small multileaf-collimator (MLC) defined beamlets. Two silver-halide films (SHFs), Kodak XV2 and EDR2, and two radiochromic films (RCFs), Gafchromic HS and MD55-2, were irradiated by MLC-defined 1x1 cm{sup 2} beamlets from a Varian 2100 C/D linac equipped with a 120-leaf MLC. The beamlets were delivered with the accelerator gantry set laterally (90 deg. rotation) upon a solid-water compression film phantom at 100 cm source-to-surface distance which was positioned with the films parallel to the beam axis. Beamlets were delivered at central axis, 5.0 cm, and 10.5 cm off-axis for both leaf-end and leaf-side defined beamlets. The film dosimetry was performed using a quantitative optical density (OD) imaging system that was validated in a previous study. No significant differences between SHF and RCF measurements were observed in percentage depth doses, horizontal depth profiles, or two-dimension spatial isodose distributions in both the central axis and off-axis measurements. We found that regardless of the type of film used, RCF or SHF, a consistent data set for small beam dose modeling was generated. Previous validation studies based on the use of RCF and OD imaging system would indicate that all film produce an accurate result for small beam characterization.

  14. Emittance dependence on anode morphology of an ion beam provided by laser ablation

    NASA Astrophysics Data System (ADS)

    Velardi, L.; Delle Side, D.; Nassisi, V.

    2014-07-01

    In this work, we studied the characteristics of ion beams generated by Platone accelerator in different anode configurations. The accelerator is a laser ion source with two gaps which accelerate the ions in cascade. The laser is a ns pulsed KrF able to apply irradiances of 109-1010 W/cm2. The target ablated was pure disk of Cu. The accelerating voltage applied in this work was 60 kV. The emittance evaluation was performed by the pepper pot method utilizing radio-chromic films, EBT Gafchromic, as sensible targets. The study was performed by varying the geometric configuration of the anode (the extracting electrode), modifying the hole morphology, e.g. a plane and curved grid were mounted in order to change the extraction configuration. The results were compared with the ones obtained with the extraction hole without any grid. For the normalized emittance the lowest value was 0.20π mm mrad.

  15. On the Emittance dependence on anode morphology of laser induced ion beams

    NASA Astrophysics Data System (ADS)

    Velardi, L.; Delle Side, D.; Nassisi, V.

    2014-04-01

    In this work, we studied the characteristics of ion beams generated by PLATONE accelerator in different anode configurations. The accelerator is a laser ion source with two gaps which accelerate the ions in cascade. The laser is an excimer KrF able to work at irradiances of 108-1010 W/cm2. The target ablated was disk of Cu. The accelerating voltage applied in this work was 60 kV. The emittance evaluation was performed by the pepper pot method utilising radio-chromic films, EBT Gafchromic, as sensible targets. The study was performed by varying the geometric configuration of the anode (the extracting electrode), modifying the hole morphology. A plane and curved grids were mounted in order to change the extraction configuration. The results were compared with the ones obtained with the extraction hole without any grid. For the normalized emittance the lowest value found was 0.20 π mm mrad.

  16. Development of Laser Plasma X-ray Microbeam Irradiation System and Radiation Biological Application

    NASA Astrophysics Data System (ADS)

    Sato, Katsutoshi; Nishikino, Masaharu; Numasaki, Hodaka; Kawachi, Tetsuya; Teshima, Teruki; Nishimura, Hiroaki

    Laser plasma x-ray source has the features such as ultra short pulse, high brilliance, monochromaticity, and focusing ability. These features are excellent compared with conventional x-ray source. In order to apply the laser plasma x-ray source to the biomedical study and to more closely research the radiobilogical responce of the cancer cell such as radiation induced bystander effect, we have developed x-ray microbeam system using laser plasma x-ray source. The absorbed dose of laser plasma x-ray was estimated with Gafchromic EBT film and DNA double strand breaks on the cells were detected by immunofluorescence staining. When the cells were irradiated with laser plasma x-ray, the circular regions existing γ-H2AX positive cells were clearly identified. The usefulness of the laser plasma x-ray on the radiobiological study was proved in this research.

  17. Film ispalators

    SciTech Connect

    Startsev, Aleksandr V; Stoilov, Yurii Yu

    2002-05-31

    New physical objects, ispalators based on free soap films, exhibit persistent flows of the soap solution in open and closed volumes in air with additions of gases of the C{sub 8}F{sub 18} type (p = 20 Torr) at temperature drops on the films of the order of tenths and hundredths of kelvin. The flows move continuously at a velocity of 5 - 20 cm s{sup -1}. It is found that the parts of an inclined ispalator film show anomalous behaviour upon heating: their weight increases and they move downward over the film, whereas the unheated parts of the film move upward. Continuous radial vortex flows accompanied by the formation and washing of the regions of a thin black film are observed on circular films in closed volumes upon their uniform external cooling by evaporating water for 5 - 10 hours. The rapid flows make film ispalators the efficient heat carriers, which operate at small temperature drops (tenths and hundredths of kelvin) and surpass copper in the amount of thermal energy being transferred. The outlook for the further study and applications of film ispalators for detecting thermal fields and laser radiation is discussed. (laser applications and other topics in quantum electronics)

  18. Respiratory gated radiotherapy-pretreatment patient specific quality assurance

    PubMed Central

    Thiyagarajan, Rajesh; Sinha, Sujit Nath; Ravichandran, Ramamoorthy; Samuvel, Kothandaraman; Yadav, Girigesh; Sigamani, Ashok Kumar; Subramani, Vikraman; Raj, N. Arunai Nambi

    2016-01-01

    Organ motions during inter-fraction and intra-fraction radiotherapy introduce errors in dose delivery, irradiating excess of normal tissue, and missing target volume. Lung and heart involuntary motions cause above inaccuracies and gated dose delivery try to overcome above effects. Present work attempts a novel method to verify dynamic dose delivery using a four-dimensional (4D) phantom. Three patients with mobile target are coached to maintain regular and reproducible breathing pattern. Appropriate intensity projection image set generated from 4D-computed tomography (4D-CT) is used for target delineation. Intensity modulated radiotherapy plans were generated on selected phase using CT simulator (Siemens AG, Germany) in conjunction with “Real-time position management” (Varian, USA) to acquire 4D-CT images. Verification plans were generated for both ion chamber and Gafchromic (EBT) film image sets. Gated verification plans were delivered on the phantom moving with patient respiratory pattern. We developed a MATLAB-based software to generate maximum intensity projection, minimum intensity projections, and average intensity projections, also a program to convert patient breathing pattern to phantom compatible format. Dynamic thorax quality assurance (QA) phantom (Computerized Imaging Reference Systems type) is used to perform the patient specific QA, which holds an ion chamber and film to measure delivered radiation intensity. Exposed EBT films are analyzed and compared with treatment planning system calculated dose. The ion chamber measured dose shows good agreement with planned dose within ± 0.5% (0.203 ± 0.57%). Gamma value evaluated from EBT film shows passing rates 92–99% (96.63 ± 3.84%) for 3% dose and 3 mm distance criteria. Respiratory gated treatment delivery accuracy is found to be within clinically acceptable level. PMID:27051173

  19. Respiratory gated radiotherapy-pretreatment patient specific quality assurance.

    PubMed

    Thiyagarajan, Rajesh; Sinha, Sujit Nath; Ravichandran, Ramamoorthy; Samuvel, Kothandaraman; Yadav, Girigesh; Sigamani, Ashok Kumar; Subramani, Vikraman; Raj, N Arunai Nambi

    2016-01-01

    Organ motions during inter-fraction and intra-fraction radiotherapy introduce errors in dose delivery, irradiating excess of normal tissue, and missing target volume. Lung and heart involuntary motions cause above inaccuracies and gated dose delivery try to overcome above effects. Present work attempts a novel method to verify dynamic dose delivery using a four-dimensional (4D) phantom. Three patients with mobile target are coached to maintain regular and reproducible breathing pattern. Appropriate intensity projection image set generated from 4D-computed tomography (4D-CT) is used for target delineation. Intensity modulated radiotherapy plans were generated on selected phase using CT simulator (Siemens AG, Germany) in conjunction with "Real-time position management" (Varian, USA) to acquire 4D-CT images. Verification plans were generated for both ion chamber and Gafchromic (EBT) film image sets. Gated verification plans were delivered on the phantom moving with patient respiratory pattern. We developed a MATLAB-based software to generate maximum intensity projection, minimum intensity projections, and average intensity projections, also a program to convert patient breathing pattern to phantom compatible format. Dynamic thorax quality assurance (QA) phantom (Computerized Imaging Reference Systems type) is used to perform the patient specific QA, which holds an ion chamber and film to measure delivered radiation intensity. Exposed EBT films are analyzed and compared with treatment planning system calculated dose. The ion chamber measured dose shows good agreement with planned dose within ± 0.5% (0.203 ± 0.57%). Gamma value evaluated from EBT film shows passing rates 92-99% (96.63 ± 3.84%) for 3% dose and 3 mm distance criteria. Respiratory gated treatment delivery accuracy is found to be within clinically acceptable level. PMID:27051173

  20. Dosimetry of biological irradiations using radiochromic films.

    PubMed

    Aldelaijan, S; Nobah, A; Alsbeih, G; Moftah, B; Aldahlawi, I; Alzahrany, A; Tomic, N; Devic, S

    2013-05-21

    Delivering accurate radiation dose to blood specimens during biological irradiations is essential in quantifying damage of ionizing radiation. To estimate dose to blood samples as accurately as possible, pieces of EBT2 model GAFCHROMIC™ film were placed within an approximately 10 mm finely ground rice layer that was used to simulate test specimens inside 40 mL plastic flasks. Irradiations of flasks were carried out using an X-RAD 320 irradiator with a beam quality of 320 kVp and a measured half value layer of 1.12 mm Cu, in air and in a full scattering setup which consisted of either rice or Solid Water™ (SW) surrounding flasks, filled to the same level at top of the flasks, together with a 5 cm thick SW slab beneath them. Outputs, per cent depth doses and beam profiles at different depths were measured and compared between setups. For the same setting, the dose delivered to the middle flask under the full scattering setup is 22% larger than with the in-air setup at the depth of the specimen and 9.2% more homogeneous across the specimen thickness of 10 mm (2.3% variation in comparison to the surface). Rice showed a fairly similar performance to SW within 1% at the same depth of 10 mm. Experimental setup based on full scattering conditions was shown to provide faster, more homogenous and fairly uniform dose delivery to biological specimens in comparison to conventionally used in-air setups. PMID:23603810

  1. SU-E-T-560: Commissioning An HDR Freiburg Flap Applicator for Skin Lesion Treatment

    SciTech Connect

    Dou, K; Li, B; Lerma, F; Aroumougame, V; Sarfaraz, M; Laser, B; Jacobs, M

    2014-06-01

    Purpose: Flexible Freiburg flap used with high dose rate afterloaders is easy to cut into any size for any body site and to dwell with a precise source position, conforms to curved skin surface and then to the planned target. However, unlike intracavity or interstitial situations, incomplete scatter environment due to flap applicators exposed to air might lead to dose difference between the delivered and planned. This research is focused on the dose deviation of incomplete scatter versus full scatter. Methods: A 12x12 cm of Freiburg flap applicator was used for the validation. A Nucletron Oncentra Brachy Ver. 4.3 treatment planning system (TPS) was used for treatment planning. However, no heterogeneity correction incorporated into the brachytherap TPS needs to be considered. A Philips Brilliance CT Big Bore was employed for CT scan. Radiation was delivered using a Nucletron HDR remote afterloader system. A 10cm bolus was used to cover the flap for obtaining a full scatter. An OSL, ion chamber, and Gafchromic EBT2 film were used for commissioning the flap applicator. Results: The applicator calibration at 5mm depth was performed using an OSL dosimeter. Applicator source dwelling positions with 1D and 2D array exposed to and recorded by Gafchromic EBT2 film showed an agreement within 1mm. 1D array of Freiburg flap exhibited 4.2% cooler in dose with incomplete scatter than full scatter. 2D array showed 7.1% lower in dose for incomplete scatter than full scatter. The deviation was found more than 10% beyond 8cm in depth. Conclusion: Significant dose deviation caused by the incomplete scatter environment was found to be 7.1% at 1cm depth. This deviation was increased with increasing depth. The inaccuracy resulted from the incomplete scatter can be fixed by either placing a bolus on the top of the flap or making the plan at least 7% hotter.

  2. About Films.

    ERIC Educational Resources Information Center

    Christman, Robert; Krockover, Gerald H.

    1984-01-01

    Lists and briefly describes 46 college-level films. Films are arranged in the following categories: volcanism and earthquakes; plate tectonics; energy, water, and environmental concerns; petroleum and coal; astronomy; space exploration, space shuttle; paleontology; geomorphology; and mineralogy, petrology, and economic geology. (BC)

  3. Thin Films

    NASA Astrophysics Data System (ADS)

    Khorshidi, Zahra; Bahari, Ali; Gholipur, Reza

    2014-11-01

    Effect of annealing temperature on the characteristics of sol-gel-driven Ta ax La(1- a) x O y thin film spin-coated on Si substrate as a high- k gate dielectric was studied. Ta ax La(1- a) x O y thin films with different amounts of a were prepared (as-prepared samples). X-ray diffraction measurements of the as-prepared samples indicated that Ta0.3 x La0.7 x Oy film had an amorphous structure. Therefore, Ta0.3 x La0.7 x O y film was chosen to continue the present studies. The morphology of Ta0.3 x La0.7 x O y films was studied using scanning electron microscopy and atomic force microscopy techniques. The obtained results showed that the size of grain boundaries on Ta0.3 x La0.7 x O y film surfaces was increased with increasing annealing temperature. Electrical and optical characterizations of the as-prepared and annealed films were investigated as a function of annealing temperature using capacitance-voltage ( C- V) and current density-voltage ( J- V) measurements and the Tauc method. The obtained results demonstrated that Ta0.3 x La0.7 x O y films had high dielectric constant (≈27), wide band gap (≈4.5 eV), and low leakage current density (≈10-6 A/cm2 at 1 V).

  4. Thin Film?

    NASA Astrophysics Data System (ADS)

    Kariper, İ. Afşin

    2014-09-01

    This study focuses on the critical surface tension of lead sulfite (PbSO3) crystalline thin film produced with chemical bath deposition on substrates (commercial glass).The PbSO3 thin films were deposited at room temperature at different deposition times. The structural properties of the films were defined and examined according to X-ray diffraction (XRD) and the XRD results such as dislocation density, average grain size, and no. of crystallites per unit area. Atomic force microscopy was used to measure the film thickness and the surface properties. The critical surface tension of the PbSO3 thin films was measured with an optical tensiometer instrument and calculated using the Zisman method. The results indicated that the critical surface tension of films changed in accordance with the average grain size and film thickness. The film thickness increased with deposition time and was inversely correlated with surface tension. The average grain size increased according to deposition time and was inversely correlated with surface tension.

  5. Film Language.

    ERIC Educational Resources Information Center

    Hudlin, Edward W.

    1979-01-01

    The author briefly surveys some of the claims made about the presumed nature of film as language and some of the problems that arise. He considers the views of two influential schools of film criticism: the Russian formalists (Pudovkin and Eisenstein) and the British semiologist (Peter Wollen). (Author/SJL)

  6. Cell irradiation setup and dosimetry for radiobiological studies at ELBE

    NASA Astrophysics Data System (ADS)

    Zeil, K.; Beyreuther, E.; Lessmann, E.; Wagner, W.; Pawelke, J.

    2009-07-01

    The radiation source ELBE delivers different types of secondary radiation, which is used for cell irradiation studies in radiobiological research. Thereby an important issue is the determination of the biological effectiveness of photon radiation as a function of photon energy by using low-energetic, monochromatic channeling radiation (10-100 keV) and high-energetic bremsstrahlung (up to 40 MV). Radiobiological studies at the research facility ELBE demand special technical and dosimetric prerequisites. Therefore, a cell irradiation system (CIS) has been designed, constructed and installed at the beam line. The CIS allows automatic irradiation of a larger cell sample number and the compensation of spatial inhomogeneity of the dose distribution within the beam spot. The recently introduced GafChromic ® EBT radiochromic film model has been used to verify the cell irradiation dose deposition achieving a dose uncertainty of <5%. Both, the installed cell irradiation system and the developed dosimetric procedure based on the use of the EBT film have been experimentally tested at ELBE. The biological effectiveness of 34 MV bremsstrahlung with respect to 200 kV X-rays from a conventional X-ray tube has been determined. An RBE value of 0.75 has been measured in good agreement with literature.

  7. Dosimetric characterization of CyberKnife radiosurgical photon beams using polymer gels

    SciTech Connect

    Pantelis, E.; Antypas, C.; Petrokokkinos, L.; Karaiskos, P.; Papagiannis, P.; Kozicki, M.; Georgiou, E.; Sakelliou, L.; Seimenis, I.

    2008-06-15

    Dose distributions registered in water equivalent, polymer gel dosimeters were used to measure the output factors and off-axis profiles of the radiosurgical photon beams employed for CyberKnife radiosurgery. Corresponding measurements were also performed using a shielded silicon diode commonly employed for CyberKnife commissioning, the PinPoint ion chamber, and Gafchromic EBT films, for reasons of comparison. Polymer gel results of this work for the output factors of the 5, 7.5, and 10 mm diameter beams are (0.702{+-}0.029), (0.872{+-}0.039), and (0.929{+-}0.041), respectively. Comparison of polymer gel and diode measurements shows that the latter overestimate output factors of the two small beams (5% for the 5 mm beam and 3% for the 7.5 mm beams). This is attributed to the nonwater equivalence of the high atomic number silicon material of the diode detector. On the other hand, the PinPoint chamber is found to underestimate output factors up to 10% for the 5 mm beam due to volume averaging effects. Polymer gel and EBT film output factor results are found in close agreement for all beam sizes, emphasizing the importance of water equivalence and fine detector sensitive volume for small field dosimetry. Relative off-axis profile results are in good agreement for all dosimeters used in this work, with noticeable differences observed only in the PinPoint estimate of the 80%-20% penumbra width, which is relatively overestimated.

  8. Film Makers On Film Making.

    ERIC Educational Resources Information Center

    Geduld, Harry M., Ed.

    This collection includes essays by and interviews with more than 30 film-makers, both classic and contemporary, on the subjects of their major interests and procedures in making films. The directors are: Louis Lumiere, Cecil Hepworth, Edwin S. Porter, Mack Sennett, David W. Griffith, Robert Flaherty, Charles Chaplin, Eric von Stroheim, Dziga…

  9. Development of radiochromic film for spatially quantitative dosimetric analysis of indirect ionizing radiation fields

    NASA Astrophysics Data System (ADS)

    Brady, Samuel Loren

    Two types of radiochromic films (RCF) were characterized for this work: EBT and XRQA film. Both films were investigated for: radiation interaction with film structure; light interaction with film structure for optimal film readout (densitometry) sensitivity; range of absorbed dose measurements; dependence of film dose measurement response as a function of changing radiation energy; fractionation and dose rate effects on film measurement response; film response sensitivity to ambient factors; and stability of measured film response with time. EBT film was shown to have the following properties: near water equivalent atomic weight (Zeff); dynamic dose range of 10 -1-102 Gy; 3% change in optical density (OD) response for a single exposure level when exposed to radiation energies from (75-18,000) kV; and best digitized using transmission densitometry. XRQA film was shown to have: a Zeff of ˜25; a 12 fold increase in sensitivity at lower photon energies for a dynamic dose range of 10-3-100 Gy, a difference of 25% in OD response when comparing 120 kV to 320 kV, and best digitized using reflective densitometry. Both XRQA and EBT films were shown to have: a temporal stability (DeltaOD) of ˜1% for t > 24 hr post film exposure for up to ˜20 days; a change in dose response of ˜0.03 mGy hr-1 when exposed to fluorescent room lighting at standard room temperature and humidity levels; a negligible dose rate and fractionation effect when operated within the optimal dose ranges; and a light wavelength dependence with dose for film readout. The flat bed scanner was chosen as the primary film digitizer due to its availability, cost, OD range, functionality (transmission and reflection scanning), and digitization speed. As a cost verses functionality comparison, the intrinsic and operational limitations were determined for two flat bed scanners. The EPSON V700 and 10000XL exhibited equal spatial and OD accuracy. The combined precision of both the scanner light sources and CCD

  10. Development of radiochromic film for spatially quantitative dosimetric analysis of indirect ionizing radiation fields

    NASA Astrophysics Data System (ADS)

    Brady, Samuel Loren

    Two types of radiochromic films (RCF) were characterized for this work: EBT and XRQA film. Both films were investigated for: radiation interaction with film structure; light interaction with film structure for optimal film readout (densitometry) sensitivity; range of absorbed dose measurements; dependence of film dose measurement response as a function of changing radiation energy; fractionation and dose rate effects on film measurement response; film response sensitivity to ambient factors; and stability of measured film response with time. EBT film was shown to have the following properties: near water equivalent atomic weight (Zeff); dynamic dose range of 10 -1-102 Gy; 3% change in optical density (OD) response for a single exposure level when exposed to radiation energies from (75-18,000) kV; and best digitized using transmission densitometry. XRQA film was shown to have: a Zeff of ˜25; a 12 fold increase in sensitivity at lower photon energies for a dynamic dose range of 10-3-100 Gy, a difference of 25% in OD response when comparing 120 kV to 320 kV, and best digitized using reflective densitometry. Both XRQA and EBT films were shown to have: a temporal stability (DeltaOD) of ˜1% for t > 24 hr post film exposure for up to ˜20 days; a change in dose response of ˜0.03 mGy hr-1 when exposed to fluorescent room lighting at standard room temperature and humidity levels; a negligible dose rate and fractionation effect when operated within the optimal dose ranges; and a light wavelength dependence with dose for film readout. The flat bed scanner was chosen as the primary film digitizer due to its availability, cost, OD range, functionality (transmission and reflection scanning), and digitization speed. As a cost verses functionality comparison, the intrinsic and operational limitations were determined for two flat bed scanners. The EPSON V700 and 10000XL exhibited equal spatial and OD accuracy. The combined precision of both the scanner light sources and CCD

  11. Piezoelectric Film.

    ERIC Educational Resources Information Center

    Garrison, Steve

    1992-01-01

    Presents activities that utilize piezoelectric film to familiarize students with fundamental principles of electricity. Describes classroom projects involving chemical sensors, microbalances, microphones, switches, infrared sensors, and power generation. (MDH)

  12. Film Reviews

    ERIC Educational Resources Information Center

    Ladd, George T.

    1974-01-01

    Briefly describes films about the following topics: water cycles, the energy crisis, the eruption of Mt. Aetna, the hot springs of Yellowstone National Park, and methods of using pine cones to determine the ages of ancient civilizations. (MLH)

  13. Polymer films

    DOEpatents

    Granick, Steve; Sukhishvili, Svetlana A.

    2004-05-25

    A film contains a first polymer having a plurality of hydrogen bond donating moieties, and a second polymer having a plurality of hydrogen bond accepting moieties. The second polymer is hydrogen bonded to the first polymer.

  14. Polymer films

    DOEpatents

    Granick, Steve; Sukhishvili, Svetlana A.

    2008-12-30

    A film contains a first polymer having a plurality of hydrogen bond donating moieties, and a second polymer having a plurality of hydrogen bond accepting moieties. The second polymer is hydrogen bonded to the first polymer.

  15. SU-E-T-533: LET Dependence Correction of Radiochromic Films for Application in Low Energy Proton Irradiation

    SciTech Connect

    Reinhardt, S; Wuerl, M; Assmann, W; Parodi, K; Greubel, C; Wilkens, J; Hillbrand, M; Mairani, A

    2015-06-15

    Purpose: Many cell irradiation experiments with low-energy laser-driven ions rely on radiochromic films (RCF), because of their dose-rate independent response and superior spatial resolution. RCF dosimetry in low-energy ion beams requires a correction of the LET dependent film response. The relative efficiency (RE), the ratio of photon to proton dose that yields the same film darkening, is a measure for the film’s LET dependence. A direct way of RE determination is RCF irradiation with low-energy mono-energetic protons and hence, well-defined LET. However, RE is usually determined using high energy proton depth dose measurements where RE corrections require knowledge of the average LET in each depth, which can be either track (tLET) or dose (dLET) averaged. The appropriate LET concept has to be applied to allow a proper film response correction. Methods: Radiochromic EBT2 and EBT3 films were irradiated in clinical photon and proton beams. For each depth of the 200 MeV proton depth dose curve, tLET and dLET were calculated by special user routines from the Monte Carlo code FLUKA. Additional irradiations with mono-energetic low energy protons (4–20 MeV) serve as reference for the RE determination. Results: The difference of dLET and tLET increases with depth, with the dLET being almost twice as large as the tLET for the maximum depth. The comparison with mono-energetic measurements shows a good agreement of the RE for the dLET concept, while a considerably steeper drop in RE is observed when applying the tLET. Conclusion: RCF can be used as reference dosimeter for biomedical experiments with low-energy proton beams if appropriate LET corrections are applied. When using depth dose measurements from clinical proton accelerators for these corrections, the concept of dLET has to be applied. Acknowledgement: This work was funded by the DFG Cluster of Excellence ‘Munich-Centre for Advanced Photonics’ (MAP). This work was funded by the DFG Cluster of Excellence

  16. SU-F-18C-10: Clinical Implementation of Radiochromic Film Based CTDI Measurements

    SciTech Connect

    Quintero, C; Seuntjens, J; Devic, S; Tomic, N; DeBlois, F

    2014-06-15

    Purpose: To evaluate accuracy and reproducibility of a radiochromic film-based protocol to measure computer tomography dose index (CTDI) as a part of annual QA on CT scanners and kV CBCT systems attached to linear accelerators. Methods: Energy dependence of Gafchromic XR-QA2(R) film model was tested over imaging beam qualities (50 – 140 kVp). Film pieces were irradiated in air to known values of air kerma in air (up to 10 cGy). Change in film reflectance was determined with an in-house written code using images produced by a flatbed document scanner. Calibration curves for each beam quality were created, and film responses for same air-kerma values were compared.Sets of film strips were placed into holes of a CTDI phantom and irradiated for several clinical scanning protocols on CT-simulators and CBCT systems. Film reflectance change was converted into dose to water and used to calculate CTDIvol values. Measured CTDIvol values were compared to tabulated CTDIvol values. Results: Average variations of ±5.2% in the mean film reflectance change were observed in the energy range of 80 to 140 keV, and 11.1% between 50 and 140 keV. The averaged measured CTDI values presented a mean variation for the same machine and protocol of 2.6%. However, measured CTDI values were in average 10% lower than tabulated CTDI values for CT-simulators, and 44% higher for CBCT systems. Conclusion: We found that in relatively broad range of beam qualities used in diagnostic radiology variation of film response is within ±5% resulting in ±15% systematic error in dose estimates if a single calibration curve is used. Relatively large discrepancy between measured and tabulated CTDI values for different protocols and imaging systems used within radiotherapy department strongly support the trend towards replacing CTDI value with equilibrium dose measurement in the center of cylindrical phantom as suggested by TG-111. This work was supported by the Natural Sciences and Engineering Research

  17. SU-E-T-276: Radiochromic Film Dosimetry: Solving the Lateral Response Puzzle

    SciTech Connect

    Lewis, D; Chan, M

    2014-06-01

    Purpose: A known factor affecting the accuracy of radiochromic film dosimetry is the lateral response artifact induced by off-center placement of film on flatbed scanners. This work reports a practical solution to eliminate such artifacts for IMRT dose QA. Methods: EBT3 film calibration strips (3.5×32cm{sup 2}) were exposed at a depth of 5cm on a Varian Trilogy to a 40×40cm{sup 2} open field at six doses between 120 and 900cGy using 6MV photons. A sheet of film was similarly exposed with a large IMRT field. Together with a strip of unexposed film from the same lot the films were digitized at different lateral positions on an Epson 10000XL scanner. The calibration strips covered nearly the full width of the scanner window. Calibration and response data were used to generate the IMRT dose map using multi-channel dosimetry. The scanner response was determined as a function of lateral position of the scanned film. The response at any lateral position L is related to the response at the center, C, of the scanner by Response(C,D) = aL+bL*Response(L,D) where D is dose and the coefficients aL and bL are constant for each lateral location. The coefficients aL and bL were determined by measuring films exposed to different doses at the center of the scanner and various lateral positions. Results: Applying this relationship to raw calibration responses, the values at any lateral position can be corrected to an equivalent response as if that film was located centrally. Our study found that calibration curves at the different lateral positions are correlated by a simple two-point re-scaling and thereby the correction could be made. Conclusion: The work reported elaborates on the process and demonstrates improvements in radiochromic film dosimetry using this simple procedure to eliminate the lateral response artifact. Acknowledgements: Supported in part by a grant from Ashland Inc. D Lewis is an employee of Ashland Inc., the manufacturer of radiochromic film; M Chan is

  18. SU-E-T-188: Film Dosimetry Verification of Monte Carlo Generated Electron Treatment Plans

    SciTech Connect

    Enright, S; Asprinio, A; Lu, L

    2014-06-01

    Purpose: The purpose of this study was to compare dose distributions from film measurements to Monte Carlo generated electron treatment plans. Irradiation with electrons offers the advantages of dose uniformity in the target volume and of minimizing the dose to deeper healthy tissue. Using the Monte Carlo algorithm will improve dose accuracy in regions with heterogeneities and irregular surfaces. Methods: Dose distributions from GafChromic{sup ™} EBT3 films were compared to dose distributions from the Electron Monte Carlo algorithm in the Eclipse{sup ™} radiotherapy treatment planning system. These measurements were obtained for 6MeV, 9MeV and 12MeV electrons at two depths. All phantoms studied were imported into Eclipse by CT scan. A 1 cm thick solid water template with holes for bonelike and lung-like plugs was used. Different configurations were used with the different plugs inserted into the holes. Configurations with solid-water plugs stacked on top of one another were also used to create an irregular surface. Results: The dose distributions measured from the film agreed with those from the Electron Monte Carlo treatment plan. Accuracy of Electron Monte Carlo algorithm was also compared to that of Pencil Beam. Dose distributions from Monte Carlo had much higher pass rates than distributions from Pencil Beam when compared to the film. The pass rate for Monte Carlo was in the 80%–99% range, where the pass rate for Pencil Beam was as low as 10.76%. Conclusion: The dose distribution from Monte Carlo agreed with the measured dose from the film. When compared to the Pencil Beam algorithm, pass rates for Monte Carlo were much higher. Monte Carlo should be used over Pencil Beam for regions with heterogeneities and irregular surfaces.

  19. SU-E-T-516: Measurement of the Absorbed Dose Rate in Water Under Reference Conditions in a CyberKnife Unit

    SciTech Connect

    Aragon-Martinez, N; Hernandez-Guzman, A; Gomez-Munoz, A; Massillon-JL, G

    2014-06-01

    Purpose: This paper aims to measure the absorbed-dose-rate in a CyberKnife unit reference-field (6cm diameter) using three ionization chambers (IC) following the new IAEA/AAPM formalism and Gafchromic film (MD-V3-55 and EBT3) protocol according to our work reported previously. Methods: The absorbed-dose-rates were measured at 90cm and 70cm SSD in a 10cmx10cm field and at 70cm SSD in a 5.4cmx5.4cm equivalent to 6cm diameter field using a linac Varian iX. All measurements were performed at 10cm depth in water. The correction factors that account for the difference between the IC response on the reference field and the CyberKnife reference field, k-(Q-msr,Q)^(f-msr,f-ref), were evaluated and Gafchromic film were calibrated using the results obtained above. Under the CyberKnife reference conditions, the factors were used to measure the absorbed-dose-rate with IC according to the new formalism and the calibrated film was irradiated in water. The film calibration curve was used to evaluate the absorbed-dose-rate in the CyberKnife unit. Results: Difference up to 2.56% is observed between dose-rate measured with IC in the reference 10cmx10cm field, depending where the chamber was calibrated, which was not reflected in the correction factor k-(Q-msr,Q)^(f-msr,f-ref ) where variations of ~0.15%-0.5% were obtained. Within measurements uncertainties, maximum difference of 1.8% on the absorbed-dose-rate in the CyberKnife reference field is observed between all IC and the films Conclusion: Absorbed-dose-rate to water was measured in a CyberKnife reference field with acceptable accuracy (combined uncertainties ~1.32%-1.73%, k=1) using three IC and films. The MD-V3-55 film as well as the new IAEA/AAPM formalism can be considered as a suitable dosimetric method to measure absorbed-dose-rate to water in small and non-standard CyberKnife fields used in clinical treatments However, the EBT3 film is not appropriated due to the high uncertainty provided (combined uncertainty ~9%, k=1

  20. Film Credits

    ERIC Educational Resources Information Center

    Borja, Rhea R.

    2006-01-01

    With the advent of easy-to-use digital technology, schools are responding to the interests of their media-savvy students by offering more courses in filmmaking. In this article, the author features different films produced by students. Among other things, she discusses the students' growing interest in filmmaking.

  1. Thin Films

    NASA Astrophysics Data System (ADS)

    Naffouti, Wafa; Nasr, Tarek Ben; Mehdi, Ahmed; Kamoun-Turki, Najoua

    2014-11-01

    Titanium dioxide (TiO2) thin films were synthesized on glass substrates by spray pyrolysis. The effect of solution flow rate on the physical properties of the films was investigated by use of x-ray diffraction (XRD), scanning electron microscopy, atomic force microscopy (AFM), and spectrophotometry techniques. XRD analysis revealed the tetragonal anatase phase of TiO2 with highly preferred (101) orientation. AFM images showed that grain size on top of TiO2 thin films depended on solution flow rate. An indirect band gap energy of 3.46 eV was determined by means of transmission and reflection measurements. The envelope method, based on the optical transmission spectrum, was used to determine film thickness and optical constants, for example real and imaginary parts of the dielectric constant, refractive index, and extinction coefficient. Ultraviolet and visible photoluminescence emission peaks were observed at room temperature. These peaks were attributed to the intrinsic emission and to the surface defect states, respectively.

  2. Black Films and Film-Makers.

    ERIC Educational Resources Information Center

    Patterson, Lindsay, Ed.

    The development of black films and the attitudes of the film industry toward black films and black actors are some of the topics examined in this anthology of essays. Section 1, "Nigger to Supernigger," contains such articles as "The Death of Rastus: Negroes in American Films" by Thomas R. Cripps and "Folk Values in a New Medium" by Alain Locke…

  3. Comparison of film measurements and Monte Carlo simulations of dose delivered with very high-energy electron beams in a polystyrene phantom

    SciTech Connect

    Bazalova-Carter, Magdalena; Liu, Michael; Palma, Bianey; Koong, Albert C.; Maxim, Peter G. E-mail: BWLoo@Stanford.edu; Loo, Billy W. E-mail: BWLoo@Stanford.edu; Dunning, Michael; McCormick, Doug; Hemsing, Erik; Nelson, Janice; Jobe, Keith; Colby, Eric; Tantawi, Sami; Dolgashev, Valery

    2015-04-15

    Purpose: To measure radiation dose in a water-equivalent medium from very high-energy electron (VHEE) beams and make comparisons to Monte Carlo (MC) simulation results. Methods: Dose in a polystyrene phantom delivered by an experimental VHEE beam line was measured with Gafchromic films for three 50 MeV and two 70 MeV Gaussian beams of 4.0–6.9 mm FWHM and compared to corresponding MC-simulated dose distributions. MC dose in the polystyrene phantom was calculated with the EGSnrc/BEAMnrc and DOSXYZnrc codes based on the experimental setup. Additionally, the effect of 2% beam energy measurement uncertainty and possible non-zero beam angular spread on MC dose distributions was evaluated. Results: MC simulated percentage depth dose (PDD) curves agreed with measurements within 4% for all beam sizes at both 50 and 70 MeV VHEE beams. Central axis PDD at 8 cm depth ranged from 14% to 19% for the 5.4–6.9 mm 50 MeV beams and it ranged from 14% to 18% for the 4.0–4.5 mm 70 MeV beams. MC simulated relative beam profiles of regularly shaped Gaussian beams evaluated at depths of 0.64 to 7.46 cm agreed with measurements to within 5%. A 2% beam energy uncertainty and 0.286° beam angular spread corresponded to a maximum 3.0% and 3.8% difference in depth dose curves of the 50 and 70 MeV electron beams, respectively. Absolute dose differences between MC simulations and film measurements of regularly shaped Gaussian beams were between 10% and 42%. Conclusions: The authors demonstrate that relative dose distributions for VHEE beams of 50–70 MeV can be measured with Gafchromic films and modeled with Monte Carlo simulations to an accuracy of 5%. The reported absolute dose differences likely caused by imperfect beam steering and subsequent charge loss revealed the importance of accurate VHEE beam control and diagnostics.

  4. Accuracy of dose measurements and calculations within and beyond heterogeneous tissues for 6 MV photon fields smaller than 4 cm produced by Cyberknife

    SciTech Connect

    Wilcox, Ellen E.; Daskalov, George M.

    2008-06-15

    For the small radiation field sizes used in stereotactic radiosurgery, lateral electronic disequilibrium and steep dose gradients exist in a large portion of these fields, requiring the use of high-resolution measurement techniques. These relatively large areas of electronic disequilibrium make accurate dosimetry as well as dose calculation more difficult, and this is exacerbated in regions of tissue heterogeneity. Tissue heterogeneity was considered insignificant in the brain where stereotactic radiosurgery was first used. However, as this technique is expanded to the head and neck and other body sites, dose calculations need to account for dose perturbations in and beyond air cavities, lung, and bone. In a previous study we have evaluated EBT Gafchromic film (International Specialty Products, Wayne, NJ) for dosimetry and characterization of the Cyberknife radiation beams and found that it was comparable to other common detectors used for small photon beams in solid water equivalent phantoms. In the present work EBT film is used to measure dose in heterogeneous slab phantoms containing lung and bone equivalent materials for the 6 MV radiation beams of diameter 7.5 to 40 mm produced by the Cyberknife (Accuray, Sunnyvale, CA). These measurements are compared to calculations done with both the clinically utilized Raytrace algorithm as well as the newly developed Monte Carlo based algorithm available on the Cyberknife treatment planning system. Within the low density material both the measurements and Monte Carlo calculations correctly model the decrease in dose produced by a loss of electronic equilibrium, whereas the Raytrace algorithm incorrectly predicts an enhancement of dose in this region. Beyond the low density material an enhancement of dose is correctly calculated by both algorithms. Within the high density bone heterogeneity the EBT film measurements represent dose to unit density tissue in bone and agree with the Monte Carlo results when corrected to dose

  5. Film and History.

    ERIC Educational Resources Information Center

    Schaber, Robin L.

    2002-01-01

    Provides an annotated bibliography of Web sites that focus on using film to teach history. Includes Web sites in five areas: (1) film and education; (2) history of cinema; (3) film and history resources; (4) film and women; and (5) film organizations. (CMK)

  6. Film: An Introduction.

    ERIC Educational Resources Information Center

    Fell, John L.

    "Understanding Film," the opening section of this book, discusses perceptions of and responses to film and the way in which experiences with and knowledge of other media affect film viewing. The second section, "Film Elements," analyzes the basic elements of film: the use of space and time, the impact of editing, sound and color, and the effects…

  7. Computer-aided analysis of star shot films for high-accuracy radiation therapy treatment units

    NASA Astrophysics Data System (ADS)

    Depuydt, Tom; Penne, Rudi; Verellen, Dirk; Hrbacek, Jan; Lang, Stephanie; Leysen, Katrien; Vandevondel, Iwein; Poels, Kenneth; Reynders, Truus; Gevaert, Thierry; Duchateau, Michael; Tournel, Koen; Boussaer, Marlies; Cosentino, Dorian; Garibaldi, Cristina; Solberg, Timothy; De Ridder, Mark

    2012-05-01

    As mechanical stability of radiation therapy treatment devices has gone beyond sub-millimeter levels, there is a rising demand for simple yet highly accurate measurement techniques to support the routine quality control of these devices. A combination of using high-resolution radiosensitive film and computer-aided analysis could provide an answer. One generally known technique is the acquisition of star shot films to determine the mechanical stability of rotations of gantries and the therapeutic beam. With computer-aided analysis, mechanical performance can be quantified as a radiation isocenter radius size. In this work, computer-aided analysis of star shot film is further refined by applying an analytical solution for the smallest intersecting circle problem, in contrast to the gradient optimization approaches used until today. An algorithm is presented and subjected to a performance test using two different types of radiosensitive film, the Kodak EDR2 radiographic film and the ISP EBT2 radiochromic film. Artificial star shots with a priori known radiation isocenter size are used to determine the systematic errors introduced by the digitization of the film and the computer analysis. The estimated uncertainty on the isocenter size measurement with the presented technique was 0.04 mm (2σ) and 0.06 mm (2σ) for radiographic and radiochromic films, respectively. As an application of the technique, a study was conducted to compare the mechanical stability of O-ring gantry systems with C-arm-based gantries. In total ten systems of five different institutions were included in this study and star shots were acquired for gantry, collimator, ring, couch rotations and gantry wobble. It was not possible to draw general conclusions about differences in mechanical performance between O-ring and C-arm gantry systems, mainly due to differences in the beam-MLC alignment procedure accuracy. Nevertheless, the best performing O-ring system in this study, a BrainLab/MHI Vero system

  8. Reference dosimetry during diagnostic CT examination using XR-QA radiochromic film model

    SciTech Connect

    Boivin, Jonathan; Tomic, Nada; Fadlallah, Bassam; DeBlois, Francois; Devic, Slobodan

    2011-09-15

    Purpose: The authors applied 2D reference dosimetry protocol for dose measurements using XR-QA radiochromic film model during diagnostic computed tomography (CT) examinations carried out on patients and humanoid Rando phantom. Methods: Response of XR-QA model GAFCHROMIC film reference dosimetry system was calibrated in terms of Air-Kerma in air. Four most commonly used CT protocols were selected on their CT scanner (GE Lightspeed VCT 64), covering three anatomical sites (head, chest, and abdomen). For each protocol, 25 patients ongoing planned diagnostic CT examination were recruited. Surface dose was measured using four or eight film strips taped on patients' skin and on Rando phantom. Film pieces were scanned prior to and after irradiation using Epson Expression 10000XL document scanner. Optical reflectance of the unexposed film piece was subtracted from exposed one to obtain final net reflectance change, which is subsequently converted to dose using previously established calibration curves. Results: The authors' measurements show that body skin dose variation has a sinusoidal pattern along the scanning axis due to the helical movement of the x-ray tube, and a comb pattern for head dose measurements due to its axial movement. Results show that the mean skin dose at anterior position for patients is (51 {+-} 6) mGy, (29 {+-} 11) mGy, (45 {+-} 13) mGy and (38 {+-} 20) mGy for head, abdomen, angio Abdomen, and chest and abdomen protocol (UP position), respectively. The obtained experimental dose length products (DLP) show higher values than CT based DLP taken from the scanner console for body protocols, but lower values for the head protocol. Internal dose measurements inside the phantom's head indicate nonuniformity of dose distribution within scanned volume. Conclusions: In this work, the authors applied an Air-Kerma in air based radiochromic film reference dosimetry protocol for in vivo skin dose measurements. In this work, they employed green channel extracted

  9. Ferroelectric ultrathin perovskite films

    DOEpatents

    Rappe, Andrew M; Kolpak, Alexie Michelle

    2013-12-10

    Disclosed herein are perovskite ferroelectric thin-film. Also disclosed are methods of controlling the properties of ferroelectric thin films. These films can be used in a variety materials and devices, such as catalysts and storage media, respectively.

  10. Dosimetric performance and array assessment of plastic scintillation detectors for stereotactic radiosurgery quality assurance

    SciTech Connect

    Gagnon, Jean-Christophe; Theriault, Dany; Guillot, Mathieu; Archambault, Louis; Beddar, Sam; Gingras, Luc; Beaulieu, Luc

    2012-01-15

    Purpose: To compare the performance of plastic scintillation detectors (PSD) for quality assurance (QA) in stereotactic radiosurgery conditions to a microion-chamber (IC), Gafchromic EBT2 films, 60 008 shielded photon diode (SD) and unshielded diodes (UD), and assess a new 2D crosshair array prototype adapted to small field dosimetry. Methods: The PSD consists of a 1 mm diameter by 1 mm long scintillating fiber (BCF-60, Saint-Gobain, Inc.) coupled to a polymethyl-methacrylate optical fiber (Eska premier, Mitsubishi Rayon Co., Ltd., Tokyo, Japan). Output factors (S{sub c,p}) for apertures used in radiosurgery ranging from 4 to 40 mm in diameter have been measured. The PSD crosshair array (PSDCA) is a water equivalent device made up of 49 PSDs contained in a 1.63 cm radius area. Dose profiles measurements were taken for radiosurgery fields using the PSDCA and were compared to other dosimeters. Moreover, a typical stereotactic radiosurgery treatment using four noncoplanar arcs was delivered on a spherical phantom in which UD, IC, or PSD was placed. Using the Xknife planning system (Integra Radionics Burlington, MA), 15 Gy was prescribed at the isocenter, where each detector was positioned. Results: Output Factors measured by the PSD have a mean difference of 1.3% with Gafchromic EBT2 when normalized to a 10 x 10 cm{sup 2} field, and 1.0% when compared with UD measurements normalized to the 35 mm diameter cone. Dose profiles taken with the PSD crosshair array agreed with other single detectors dose profiles in spite of the presence of the 49 PSDs. Gamma values comparing 1D dose profiles obtained with PSD crosshair array with Gafchromic EBT2 and UD measured profiles shows 98.3% and 100.0%, respectively, of detector passing the gamma acceptance criteria of 0.3 mm and 2%. The dose measured by the PSD for a complete stereotactic radiosurgery treatment is comparable to the planned dose corrected for its SD-based S{sub c,p} within 1.4% and 0.7% for 5 and 35 mm diameter cone

  11. Development of a 3-dimensional dosimetry system for Leksell Gamma Knife Perfexion

    NASA Astrophysics Data System (ADS)

    Yoon, KyoungJun; Kwak, JungWon; Lee, DoHeui; Cho, ByungChul; Lee, SangWook; Ahn, SeungDo

    2015-07-01

    The purpose of our study is to develop a new, 3-dimensional dosimetry system to verify the accuracy of dose deliveries in Leksell Gamma Knife Perfexion (LGKP) (Elekta, Norcross, GA, USA). The instrument consists of a moving head phantom, an embedded thin active layer and a CCD camera system and was designed to be mounted to LGKP. As an active material concentrically located in the hemispheric head phantom, we choose Gafchromic EBT3 films and Gd2O2S:Tb phosphor sheets for dosimetric measurements. Also, to compensate for the lack of backscatter, we located a 1-cm-thick poly methyl methacrylate (PMMA) plate downstream of the active layer. The PMMA plate was transparent to scintillation light to reach the CCD with 1200 × 1200 pixels and a 5.2 µm pitch. With this system, 300 images with a 0.2-mm slice gap were acquired under each of three collimator setups, i.e. 4-mm, 8-mm, and 16-mm, respectively. The 2D projected images taken by the CCD camera were compared with the dose distributions measured by the EBT3 films under the same conditions. All 2D distributions were normalized to the maximum values derived by fitting peaks for each collimator setup. The differences in the full widths at half maximum (FWHM) of 2D profiles between CCD images and film doses were measured to be less than 0.3-mm. The scanning task for all peak regions took less than three minutes with the new instrument. So it can be utilized as a QA tool for the Gamma knife radiosurgery system instead of film dosimetry, the use of which requires much more time and many more resources.

  12. TU-F-BRE-08: Significant Variations in Measured Small Cone Output Factor for FFF Beams

    SciTech Connect

    Sudhyadhom, A; Ma, L; Kirby, N

    2014-06-15

    Purpose: To evaluate the measurement accuracy of several dosimeters for small cone output factors in two SRS/SBRT dedicated systems with Flattening Filter Free (FFF) beams: a Varian TrueBeam STx (TB) and an Accuray CyberKnife VSI (CK). Output factors (OFs) were measured for both machines and for CK, compared against a Monte Carlo model. Methods: Dose measurements were taken using three different FFF beams (TB 6XFFF, TB 10XFFF, and CK 6XFFF). Three commonly used types of dosimeters were examined in this work: a micro-ion chamber (Exradin A16), two shielded diodes (PTW TN60008 and PTW TN60017), and radiochromic film (Gafchromic EBT2). Measured OFs from these dosimeters were compared with each other and OFs measured with an Exradin W1 scintillator. Monte Carlo determined correction factors for the CK beam for the micro-ion chamber and diodes were applied to the respective OF measurements and compared against scintillator measured OFs corrected for volume averaging. Results: OFs measured for the smallest fields using the micro-ion chamber, diodes, scintillator, and film varied substantially (with up to a 16% difference between dosimeters). Micro-ion chamber and film OF measurements were up to 9% and 10%, respectively, lower than scintillator measurements for the smallest fields. OF measurements by diode were up to 6% greater than scintillator measurements for the smallest fields. With correction factors, the micro-ion chamber and diode measured OFs showed good agreement with scintillator measured OFs for the CK 6XFFF beam (within 3% and 1.5%, respectively). Conclusion: Uncorrected small field OFs vary significantly with dosimeter. The accuracy of scintillator measurements for small field OFs may be greater than the other dosimeters studied in this work (when uncorrected). Measurements involving EBT2 film may Result in lower accuracy for smaller fields (less than 10mm). Care should be taken in the choice of the dosimeter used for small field OF measurements.

  13. Contemporary Films' Mini Course on Film Study.

    ERIC Educational Resources Information Center

    Schillaci, Peter

    This minicourse on film study can be a unit in English, in arts, or in the humanities. It can help to launch a film study course or complement an introduction to theater. Whatever form it takes, it helps to build a bridge to the student's media environment. Part one, the language of images, utilizes four films which demonstrate the basic elements…

  14. RAYS: a geometrical optics code for EBT

    SciTech Connect

    Batchelor, D.B.; Goldfinger, R.C.

    1982-04-01

    The theory, structure, and operation of the code are described. Mathematical details of equilibrium subroutiones for slab, bumpy torus, and tokamak plasma geometry are presented. Wave dispersion and absorption subroutines are presented for frequencies ranging from ion cyclotron frequency to electron cyclotron frequency. Graphics postprocessors for RAYS output data are also described.

  15. [Films: China and Japan].

    ERIC Educational Resources Information Center

    Gumport, Roberta H.

    The history of filmmaking in China and Japan and film usage in teaching are considered in this document. Pointing out how films describe historical context and culture, the document also describes various techniques of film making. Films in China were heavily influenced by western models and have tended to be tools of the power structure, as…

  16. Carbon thin film thermometry

    NASA Technical Reports Server (NTRS)

    Collier, R. S.; Sparks, L. L.; Strobridge, T. R.

    1973-01-01

    The work concerning carbon thin film thermometry is reported. Optimum film deposition parameters were sought on an empirical basis for maximum stability of the films. One hundred films were fabricated for use at the Marshall Space Flight Center; 10 of these films were given a precise quasi-continuous calibration of temperature vs. resistance with 22 intervals between 5 and 80 K using primary platinum and germanium thermometers. Sensitivity curves were established and the remaining 90 films were given a three point calibration and fitted to the established sensitivity curves. Hydrogen gas-liquid discrimination set points are given for each film.

  17. Hydrogen film cooling investigation

    NASA Technical Reports Server (NTRS)

    Rousar, D. C.; Ewen, R. L.

    1973-01-01

    Effects of flow turning, flow acceleration, and supersonic flow on film cooling were determined experimentally and correlated in terms of an entrainment film cooling model. Experiments were conducted using thin walled metal test sections, hot nitrogen mainstream gas, and ambient hydrogen or nitrogen as film coolants. The entrainment film cooling model relates film cooling effectiveness to the amount of mainstream gases entrained with the film coolant in a mixing layer. The experimental apparatus and the analytical model used are described in detail and correlations for the entrainment fraction and film coolant-to-wall heat transfer coefficient are presented.

  18. Iterative reconstruction of volumetric modulated arc radiotherapy plans using control point basis vectors

    NASA Astrophysics Data System (ADS)

    Barbiere, Joseph C.; Kapulsky, Alexander; Ndlovu, Alois

    2014-03-01

    Volumetric Modulated Arc Radiotherapy is an innovative technique currently utilized to efficiently deliver complex treatments. Dose rate, speed of rotation, and field shape are continuously varied as the radiation source rotates about the patient. Patient specific quality assurance is performed to verify that the delivered dose distribution is consistent with the plan formulated in a treatment planning system. The purpose of this work is to present novel methodology using a Gafchromic EBT3 film image of a patient plan in a cylindrical phantom and calculating the delivered MU per control point. Images of two dimensional plan dose matrices and film scans are analyzed using MATLAB with the imaging toolbox. Dose profiles in a ring corresponding to the film position are extracted from the plan matrices for comparison with the corresponding measured film dose. The plan is made up of a series of individual static Control Points. If we consider these Control Points a set of basis vectors, then variations in the plan can be represented as the weighted sum of the basis. The weighing coefficients representing the actual delivered MU can be determined by any available optimization tool, such as downhill simplex or non-linear programming. In essence we reconstruct an image of the delivered dose. Clinical quality assurance is performed with this technique by computing a patient plan with the measured monitor units and standard plan evaluation tools such as Dose Volume Histograms. Testing of the algorithm with known changes in the reference images indicated a correlation coefficient greater than 0.99.

  19. Proof of principle study of the use of a CMOS active pixel sensor for proton radiography

    SciTech Connect

    Seco, Joao; Depauw, Nicolas

    2011-02-15

    Purpose: Proof of principle study of the use of a CMOS active pixel sensor (APS) in producing proton radiographic images using the proton beam at the Massachusetts General Hospital (MGH). Methods: A CMOS APS, previously tested for use in s-ray radiation therapy applications, was used for proton beam radiographic imaging at the MGH. Two different setups were used as a proof of principle that CMOS can be used as proton imaging device: (i) a pen with two metal screws to assess spatial resolution of the CMOS and (ii) a phantom with lung tissue, bone tissue, and water to assess tissue contrast of the CMOS. The sensor was then traversed by a double scattered monoenergetic proton beam at 117 MeV, and the energy deposition inside the detector was recorded to assess its energy response. Conventional x-ray images with similar setup at voltages of 70 kVp and proton images using commercial Gafchromic EBT 2 and Kodak X-Omat V films were also taken for comparison purposes. Results: Images were successfully acquired and compared to x-ray kVp and proton EBT2/X-Omat film images. The spatial resolution of the CMOS detector image is subjectively comparable to the EBT2 and Kodak X-Omat V film images obtained at the same object-detector distance. X-rays have apparent higher spatial resolution than the CMOS. However, further studies with different commercial films using proton beam irradiation demonstrate that the distance of the detector to the object is important to the amount of proton scatter contributing to the proton image. Proton images obtained with films at different distances from the source indicate that proton scatter significantly affects the CMOS image quality. Conclusion: Proton radiographic images were successfully acquired at MGH using a CMOS active pixel sensor detector. The CMOS demonstrated spatial resolution subjectively comparable to films at the same object-detector distance. Further work will be done in order to establish the spatial and energy resolution of the

  20. Use of novel fibre-coupled radioluminescence and RADPOS dosimetry systems for total scatter factor measurements in small fields

    NASA Astrophysics Data System (ADS)

    Ploquin, N.; Kertzscher, G.; Vandervoort, E.; Cygler, J. E.; Andersen, C. E.; Francescon, P.

    2015-01-01

    A dosimetry system based on Al2O3:C radioluminescence (RL), and RADPOS, a novel 4D dosimetry system using microMOSFETs, were used to measure total scatter factors, ≤ft({{S}c,p}\\right)\\text{det}{{f\\text{clin}}}, for the CyberKnife robotic radiosugery system. New Monte Carlo calculated correction factors are presented and applied for the RL detector response for the 5, 7.5 and 10 mm collimators in order to correct for the detector geometry and increased photoelectric cross section of Al2O3:C relative to water. For comparison, measurements were also carried out using a micro MOSFET, PTW60012 diode and GAFCHROMIC® film (EBT and EBT2). Uncorrected ≤ft({{S}c,p}\\right)\\text{det}{{f\\text{clin}}}, were obtained by taking the ratio of the detector response for each collimator to that for the 60 mm diameter reference field. Published Monte Carlo calculated correction factors were applied to the RADPOS, microMOSFET and diode detector measurements to yield corrected field factors, Ω {{Q\\text{clin}},{{Q}\\text{msr}}}{{f\\text{clin}},{{f}\\text{msr}}}, following the terminology of a recent formalism introduced for small and composite field relative dosimetry. With corrections, the RL measured Ω {{Q\\text{clin}},{{Q}\\text{msr}}}{{f\\text{clin}},{{f}\\text{msr}}}, were 0.656  ±  0.002, 0.815  ±  0.002 and 0.865  ±  0.003 for the 5, 7.5 and 10 mm collimators, respectively. This was in good agreement with RADPOS corrected field factors of 0.650  ±  0.010, 0.816  ±  0.024 and 0.867  ±  0.010 for the 5, 7.5 and 10 mm collimators, respectively. Both RL and RADPOS total scatter factors agreed within approximately two standard deviations of the GAFCHROMIC film values (average of EBT and EBT2) of 0.640  ±  0.006, 0.806  ±  0.007 and 0.859  ±  0.09. Corrected total scatter factors for all dosimetry systems agreed within one standard deviation for collimator sizes 10-60 mm. Our study suggests that the micro

  1. Educational Films and the Historian

    ERIC Educational Resources Information Center

    Youngs, J. William T., Jr.

    1975-01-01

    Suggestions on how historians might use educational films more efficiently in the classroom are given. The range and diversity of educational films, educational film bibliography and criticism, and the classroom uses of films are discussed. (JR)

  2. Contact radiotherapy using a 50 kV X-ray system: Evaluation of relative dose distribution with the Monte Carlo code PENELOPE and comparison with measurements

    NASA Astrophysics Data System (ADS)

    Croce, Olivier; Hachem, Sabet; Franchisseur, Eric; Marcié, Serge; Gérard, Jean-Pierre; Bordy, Jean-Marc

    2012-06-01

    This paper presents a dosimetric study concerning the system named "Papillon 50" used in the department of radiotherapy of the Centre Antoine-Lacassagne, Nice, France. The machine provides a 50 kVp X-ray beam, currently used to treat rectal cancers. The system can be mounted with various applicators of different diameters or shapes. These applicators can be fixed over the main rod tube of the unit in order to deliver the prescribed absorbed dose into the tumor with an optimal distribution. We have analyzed depth dose curves and dose profiles for the naked tube and for a set of three applicators. Dose measurements were made with an ionization chamber (PTW type 23342) and Gafchromic films (EBT2). We have also compared the measurements with simulations performed using the Monte Carlo code PENELOPE. Simulations were performed with a detailed geometrical description of the experimental setup and with enough statistics. Results of simulations are made in accordance with experimental measurements and provide an accurate evaluation of the dose delivered. The depths of the 50% isodose in water for the various applicators are 4.0, 6.0, 6.6 and 7.1 mm. The Monte Carlo PENELOPE simulations are in accordance with the measurements for a 50 kV X-ray system. Simulations are able to confirm the measurements provided by Gafchromic films or ionization chambers. Results also demonstrate that Monte Carlo simulations could be helpful to validate the future applicators designed for other localizations such as breast or skin cancers. Furthermore, Monte Carlo simulations could be a reliable alternative for a rapid evaluation of the dose delivered by such a system that uses multiple designs of applicators.

  3. Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE®/optical-CT 3D dosimetry system

    PubMed Central

    Clift, Corey; Thomas, Andrew; Adamovics, John; Chang, Zheng; Das, Indra; Oldham, Mark

    2010-01-01

    Achieving accurate small field dosimetry is challenging. This study investigates the utility of a radiochromic plastic PRESAGE® read with optical-CT for the acquisition of radiosurgery field commissioning data from a Novalis Tx system with a high-definition multileaf collimator (HDMLC). Total scatter factors (Sc, p), beam profiles, and penumbrae were measured for five different radiosurgery fields (5, 10, 20, 30 and 40 mm) using a commercially available optical-CT scanner (OCTOPUS, MGS Research). The percent depth dose (PDD), beam profile and penumbra of the 10 mm field were also measured using a higher resolution in-house prototype CCD-based scanner. Gafchromic EBT® film was used for independent verification. Measurements of Sc, p made with PRESAGE® and film agreed with mini-ion chamber commissioning data to within 4% for every field (range 0.2–3.6% for PRESAGE®, and 1.6–3.6% for EBT). PDD, beam profile and penumbra measurements made with the two PRESAGE®/optical-CT systems and film showed good agreement with the high-resolution diode commissioning measurements with a competitive resolution (0.5 mm pixels). The in-house prototype optical-CT scanner allowed much finer resolution compared with previous applications of PRESAGE®. The advantages of the PRESAGE® system for small field dosimetry include 3D measurements, negligible volume averaging, directional insensitivity, an absence of beam perturbations, energy and dose rate independence. PMID:20134082

  4. Thin film metrology.

    PubMed

    Nitsch, Gerald; Flinn, Gregory

    2007-10-01

    Thin film metrology is suitable for characterising and performing quality control of a variety of coatings and films used in medical applications. The capabilities of today's systems are described. PMID:18078184

  5. SU-E-T-231: Measurements of Gold Nanoparticle-Mediated Proton Dose Enhancement Due to Particle-Induced X-Ray Emission and Activation Products Using Radiochromic Films and CdTe Detector

    SciTech Connect

    Cho, J; Cho, S; Manohar, N; Krishnan, S

    2014-06-01

    Purpose: There have been several reports of enhanced cell-killing and tumor regression when tumor cells and mouse tumors were loaded with gold nanoparticles (GNPs) prior to proton irradiation. While particle-induced xray emission (PIXE), Auger electrons, secondary electrons, free radicals, and biological effects have been suggested as potential mechanisms responsible for the observed GNP-mediated dose enhancement/radiosensitization, there is a lack of quantitative analysis regarding the contribution from each mechanism. Here, we report our experimental effort to quantify some of these effects. Methods: 5-cm-long cylindrical plastic vials were filled with 1.8 mL of either water or water mixed with cylindrical GNPs at the same gold concentration (0.3 mg Au/g) as used in previous animal studies. A piece of EBT2 radiochromic film (30-µm active-layer sandwiched between 80/175-µm outer-layers) was inserted along the long axis of each vial and used to measure dose enhancement due to PIXE from GNPs. Vials were placed at center-of-modulation (COM) and 3-cm up-/down-stream from COM and irradiated with 5 different doses (2–10 Gy) using 10-cm-SOBP 160-MeV protons. After irradiation, films were cleaned and read to determine the delivered dose. A vial containing spherical GNPs (20 mg Au/g) was also irradiated, and gamma-rays from activation products were measured using a cadmium-telluride (CdTe) detector. Results: Film measurements showed no significant dose enhancement beyond the experimental uncertainty (∼2%). There was a detectable activation product from GNPs, but it appeared to contribute to dose enhancement minimally (<0.01%). Conclusion: Considering the composition of EBT2 film, it can be inferred that gold characteristic x-rays from PIXE and their secondary electrons make insignificant contribution to dose enhancement. The current investigation also suggests negligible dose enhancement due to activation products. Thus, previously-reported GNP-mediated proton dose

  6. Film as Film; Understanding and Judging Movies.

    ERIC Educational Resources Information Center

    Perkins, V. F.

    The criteria for judging movies which are presented here are based on the belief that film criticism becomes rational, if not "objective", when it displays and inspects the nature of its evidence and the bases of its arguments. The author dissents from the view of early film theorists that montage is the essence of cinema, and that cinema is to be…

  7. Exploring the Film.

    ERIC Educational Resources Information Center

    Kuhns, William; Stanley, Robert

    The purpose of film study is defined here in the words of D. W. Griffiths: "My goal is above all to make you see." This book is intended to be used as a text in a film study course. It traces the development of films from a scientific curiosity through silent films to modern wide screen productions. A comic strip is used to demonstrate the effect…

  8. Focus on Shakespearean Films.

    ERIC Educational Resources Information Center

    Eckert, Charles W., Ed.

    This is an anthology of reviews and critical pieces of the significant and available Shakespearean films made between 1935 and 1966. Included are three general essays on Shakespearean film by Ian Johnson, Henri Lemaitre, and Geoffrey Reeves. The specific films and their reviewers are: A Midsummer's Night Dream (1935) Allardyce Nicoll and Richard…

  9. Marine Science Film Catalogue.

    ERIC Educational Resources Information Center

    Chapman, Frank L.

    Forty-eight motion picture films and filmstrips in the field of marine science are catalogued in this booklet. Following the alphabetical index, one page is devoted to each film indicating its type, producer, recommended grade level, running time, and presence of color and/or sound. A summary of film content, possible uses, and outstanding…

  10. Thick Film Interference.

    ERIC Educational Resources Information Center

    Trefil, James

    1983-01-01

    Discusses why interference effects cannot be seen with a thick film, starting with a review of the origin of interference patterns in thin films. Considers properties of materials in films, properties of the light source, and the nature of light. (JN)

  11. TEACHING COMPOSITION WITH FILM.

    ERIC Educational Resources Information Center

    COURSEN, HERBERT R., JR.

    A COMPOSITION PROGRAM DESIGNED TO GIVE UPWARD BOUND STUDENTS A FEELING OF SUCCESS WAS BASED ON FILMS WHICH THE STUDENTS VIEWED, DISCUSSED, AND WROTE ABOUT. THE FILMS FELL ROUGHLY INTO THE CATEGORIES OF SOCIAL PROBLEMS, POLITICS AND PROPAGANDA, AND ART AND MUSIC. FOLLOWING CLASS DISCUSSIONS, STUDENTS WERE REQUIRED MERELY TO "WRITE ABOUT THE FILM."…

  12. Films for Special Education.

    ERIC Educational Resources Information Center

    Tringo, John L.

    Listed are approximately 500 films in the various areas of special education. The listing has been compiled from film company catalogs, social agency listings, and National Audiovisual Center listings. Information stated is that which was in the source materials (usually name, length, source, and brief description). Films are listed under the…

  13. 99 Films on Drugs.

    ERIC Educational Resources Information Center

    Weber, David O., Ed.

    This catalog describes and evaluates 16-millimeter films about various aspects of drug use. Among the subjects covered by the 99 films are the composition and effects of different drugs, reasons why people use drugs, life in the drug culture, the problem of law enforcement, and various means of dealing with drug users. Each film is synopsized. Two…

  14. The Sponsored Film.

    ERIC Educational Resources Information Center

    Klein, Walter J.

    For public relations professionals and would-be sponsors of films, this book provides guidelines for understanding the film medium and its potential as a persuasive force in industry, government, organizations, and religious orders. For filmmakers, it brings together practical information needed to survive in the sponsored-film industry and to…

  15. Health Careers Film Guide.

    ERIC Educational Resources Information Center

    National Institutes of Health (DHEW), Bethesda, MD. Bureau of Health Manpower Education.

    This document, which represents a survey of the entire health career film field, was designed to provide information for people interested in a health career. The guide indicates that a major criteria for film selection was recency; however, some older films that give a fairly accurate image of a profession were included, with some emphasis given…

  16. Particle film technology

    Technology Transfer Automated Retrieval System (TEKTRAN)

    Particle Film Technology involves establishing a mineral particle film on the surface of a plant or plant product that: (1) is chemically inert, (2) has a mean particle diameter < 2 um, (3) is formulated to spread and create a uniform film, (4) does not physically disrupt gas exchange from the le...

  17. Getting into Film.

    ERIC Educational Resources Information Center

    London, Mel

    This book describes the various aspects of the film industry and the many jobs related to filmmaking, stressing that no "formula" exists for finding a successful career in the film industry. Chapters provide information on production, writing for film, cinematography, editing, music, sound, animation and graphics, acting and modeling, the "unsung…

  18. Poster — Thur Eve — 58: Dosimetric validation of electronic compensation for radiotherapy treatment planning

    SciTech Connect

    Gräfe, James; Khan, Rao; Meyer, Tyler

    2014-08-15

    In this study we investigate the deliverability of dosimetric plans generated by the irregular surface compensator (ISCOMP) algorithm for 6 MV photon beams in Eclipse (Varian Medical System, CA). In contrast to physical tissue compensation, the electronic ISCOMP uses MLCs to dynamically modulate the fluence of a photon beam in order to deliver a uniform dose at a user defined plane in tissue. This method can be used to shield critical organs that are located within the treatment portal or improve dose uniformity by tissue compensation in inhomogeneous regions. Three site specific plans and a set of test fields were evaluated using the γ-metric of 3%/ 3 mm on Varian EPID, MapCHECK, and Gafchromic EBT3 film with a clinical tolerance of >95% passing rates. Point dose measurements with an NRCC calibrated ionization chamber were also performed to verify the absolute dose delivered. In all cases the MapCHECK measured plans met the gamma criteria. The mean passing rate for the six EBT3 film field measurements was 96.2%, with only two fields at 93.4 and 94.0% passing rates. The EPID plans passed for fields encompassing the central ∼10 × 10 cm{sup 2} region of the detector; however for larger fields and greater off-axis distances discrepancies were observed and attributed to the profile corrections and modeling of backscatter in the portal dose calculation. The magnitude of the average percentage difference for 21 ion chamber point dose measurements and 17 different fields was 1.4 ± 0.9%, and the maximum percentage difference was −3.3%. These measurements qualify the algorithm for routine clinical use subject to the same pre-treatment patient specific QA as IMRT.

  19. Pyrolyzed thin film carbon

    NASA Technical Reports Server (NTRS)

    Tai, Yu-Chong (Inventor); Liger, Matthieu (Inventor); Harder, Theodore (Inventor); Konishi, Satoshi (Inventor); Miserendino, Scott (Inventor)

    2010-01-01

    A method of making carbon thin films comprises depositing a catalyst on a substrate, depositing a hydrocarbon in contact with the catalyst and pyrolyzing the hydrocarbon. A method of controlling a carbon thin film density comprises etching a cavity into a substrate, depositing a hydrocarbon into the cavity, and pyrolyzing the hydrocarbon while in the cavity to form a carbon thin film. Controlling a carbon thin film density is achieved by changing the volume of the cavity. Methods of making carbon containing patterned structures are also provided. Carbon thin films and carbon containing patterned structures can be used in NEMS, MEMS, liquid chromatography, and sensor devices.

  20. UV actinometer film

    NASA Technical Reports Server (NTRS)

    Coulbert, C. D.; Gupta, A.; Pitts, J.

    1980-01-01

    Cumulative UV radiation can be measured by low-cost polymer film that is unaffacted by visible light. Useful for virtually any surface, film can help paint and plastics manufacturers determine how well their products stand up against UV radiation. Actinometer film uses photochemically sensitive compound that changes its chemical composition in response to solar radiation. Extent of chemical conversion depends on length exposure and can be measured by examining film sample with spectrophotometer. Film can be exposed from several seconds up to month.

  1. Chronicles of foam films.

    PubMed

    Gochev, G; Platikanov, D; Miller, R

    2016-07-01

    The history of the scientific research on foam films, traditionally known as soap films, dates back to as early as the late 17th century when Boyle and Hooke paid special attention to the colours of soap bubbles. Their inspiration was transferred to Newton, who began systematic study of the science of foam films. Over the next centuries, a number of scientists dealt with the open questions of the drainage, stability and thickness of foam films. The significant contributions of Plateau and Gibbs in the middle/late 19th century are particularly recognized. After the "colours" method of Newton, Reinold and Rücker as well as Johhonnot developed optical methods for measuring the thickness of the thinner "non-colour" films (first order black) that are still in use today. At the beginning of the 20th century, various aspects of the foam film science were elucidated by the works of Dewar and Perrin and later by Mysels. Undoubtedly, the introduction of the disjoining pressure by Derjaguin and the manifestation of the DLVO theory in describing the film stability are considered as milestones in the theoretical development of foam films. The study of foam films gained momentum with the introduction of the microscopic foam film methodology by Scheludko and Exerowa, which is widely used today. This historical perspective serves as a guide through the chronological development of knowledge on foam films achieved over several centuries. PMID:26361708

  2. Depositing Diamondlike Carbon Films

    NASA Technical Reports Server (NTRS)

    Mirtich, M. J.; Sovey, J. S.; Banks, B. A.

    1986-01-01

    New process demonstrated to make thin films (usually thousands of angstroms to few microns thick) that have properties of diamonds. Various plasma and ion-beam techniques employed to generate films. Films made by radio-frequency plasma decomposition of hydrocarbon gas or other alkanes, by low-energy carbon-ion-beam deposition, or by ion plating and dual ion technique using carbon target. Advantages of new process over others are films produced, though amorphous, are clear, extremely hard, chemically inert, of high resistivity, and have index of refraction of 3.2 properties similar to those of single-crystal diamonds. Films have possible uses in microelectronic applications, high-energy-laser and plastic windows, corrosion protection for metals, and other applications where desired properties of film shaped during the film-formation process.

  3. Ceramic Composite Thin Films

    NASA Technical Reports Server (NTRS)

    Ruoff, Rodney S. (Inventor); Stankovich, Sasha (Inventor); Dikin, Dmitriy A. (Inventor); Nguyen, SonBinh T. (Inventor)

    2013-01-01

    A ceramic composite thin film or layer includes individual graphene oxide and/or electrically conductive graphene sheets dispersed in a ceramic (e.g. silica) matrix. The thin film or layer can be electrically conductive film or layer depending the amount of graphene sheets present. The composite films or layers are transparent, chemically inert and compatible with both glass and hydrophilic SiOx/silicon substrates. The composite film or layer can be produced by making a suspension of graphene oxide sheet fragments, introducing a silica-precursor or silica to the suspension to form a sol, depositing the sol on a substrate as thin film or layer, at least partially reducing the graphene oxide sheets to conductive graphene sheets, and thermally consolidating the thin film or layer to form a silica matrix in which the graphene oxide and/or graphene sheets are dispersed.

  4. Optical computed tomography of radiochromic gels for accurate three-dimensional dosimetry

    NASA Astrophysics Data System (ADS)

    Babic, Steven

    In this thesis, three-dimensional (3-D) radiochromic Ferrous Xylenol-orange (FX) and Leuco Crystal Violet (LCV) micelles gels were imaged by laser and cone-beam (Vista(TM)) optical computed tomography (CT) scanners. The objective was to develop optical CT of radiochromic gels for accurate 3-D dosimetry of intensity-modulated radiation therapy (IMRT) and small field techniques used in modern radiotherapy. First, the cause of a threshold dose response in FX gel dosimeters when scanned with a yellow light source was determined. This effect stems from a spectral sensitivity to multiple chemical complexes that are at different dose levels between ferric ions and xylenol-orange. To negate the threshold dose, an initial concentration of ferric ions is needed in order to shift the chemical equilibrium so that additional dose results in a linear production of a coloured complex that preferentially absorbs at longer wavelengths. Second, a low diffusion leuco-based radiochromic gel consisting of Triton X-100 micelles was developed. The diffusion coefficient of the LCV micelle gel was found to be minimal (0.036 + 0.001 mm2 hr-1 ). Although a dosimetric characterization revealed a reduced sensitivity to radiation, this was offset by a lower auto-oxidation rate and base optical density, higher melting point and no spectral sensitivity. Third, the Radiological Physics Centre (RPC) head-and-neck IMRT protocol was extended to 3-D dose verification using laser and cone-beam (Vista(TM)) optical CT scans of FX gels. Both optical systems yielded comparable measured dose distributions in high-dose regions and low gradients. The FX gel dosimetry results were crossed checked against independent thermoluminescent dosimeter and GAFChromicRTM EBT film measurements made by the RPC. It was shown that optical CT scanned FX gels can be used for accurate IMRT dose verification in 3-D. Finally, corrections for FX gel diffusion and scattered stray light in the Vista(TM) scanner were developed to

  5. SU-E-CAMPUS-T-05: Preliminary Results On a 2D Dosimetry System Based On the Optically Stimulated Luminescence of Al2O3

    SciTech Connect

    Ahmed, M; Eller, S; Yukihara, E; Schnell, E; Ahmad, S; Akselrod, M; Hanson, O

    2014-06-15

    Purpose: To develop a precise 2D dose mapping technique based on the optically stimulated luminescence (OSL) from Al{sub 2}O{sub 3} films for medical applications. Methods: A 2D laser scanning reader was developed using fast F{sup +}-center (lifetime of <7 ns) and slow F-center (lifetime of 35 ms) OSL emission from newly developed Al{sub 2}O{sub 3} films (Landauer Inc.). An algorithm was developed to correct images for both material and system properties. Since greater contribution of the F??-center emission in the recorded signal increases the readout efficiency and robustness of image corrections, Al{sub 2}O{sub 3}:C,Mg film samples are being investigated in addition to Al{sub 2}O{sub 3}:C samples. Preliminary investigations include exposure of the films to a 6 MV photon beam at 10 cm depth in solid water phantom with an SSD of 100 cm, using a 10 cm × 10 cm flat field or a 4 cm × 4 cm field with a 60° wedge filter. Kodak EDR2 radiographic film and EBT2 Gafchromic film were also exposed for comparison. Results: The results indicate that the algorithm is able to correct images and calculate 2D dose. For the wedge field irradiation, the calculated dose at the center of the field was 0.9 Gy for Al{sub 2}O{sub 3}:C and 0.87 Gy for Al{sub 2}O{sub 3}:C,Mg, whereas, the delivered dose was 0.95 Gy. A good qualitative agreement of the dose profiles was obtained between the OSL films and EDR2 and EBT2 films. Laboratory tests using a beta source suggest that a large dynamic range (10{sup −2}−10{sup 2} Gy) can be achieved using this technique. Conclusion: A 2D dosimetry system and an in-house image correction algorithm were developed for 2D film dosimetry in medical applications. The system is in the preliminary stage of development, but the data demonstrates the feasibility of this approach. This work was supported by Landauer, Inc.

  6. Documentary Elements in Early Films.

    ERIC Educational Resources Information Center

    Sanderson, Richard A.

    Focusing on documentary elements, this study examines the film content and film techniques of 681 motion pictures produced in the United States prior to 1904. Analysis of films by type, subject matter, and trends in subject matter shows that one-third of the early films are documentary in type and three-fourths of the films use subject matter of a…

  7. Coating of plasma polymerized film

    NASA Technical Reports Server (NTRS)

    Morita, S.; Ishibashi, S.

    1980-01-01

    Plasma polymerized thin film coating and the use of other coatings is suggested for passivation film, thin film used for conducting light, and solid body lubrication film of dielectrics of ultra insulators for electrical conduction, electron accessories, etc. The special features of flow discharge development and the polymerized film growth mechanism are discussed.

  8. Film Music. Factfile No. 8.

    ERIC Educational Resources Information Center

    Elsas, Diana, Ed.; And Others

    Organizations listed here with descriptive information include film music clubs and music guilds and associations. These are followed by a representative list of schools offering film music and/or film sound courses. Sources are listed for soundtrack recordings, sound effects/production music, films on film music, and oral history programs. The…

  9. Books on Film and Filmmaking.

    ERIC Educational Resources Information Center

    Ruth, Deborah D.

    This annotated bibliography of books on film and filmmaking contains references on materials related to animation production, general reference materials, student film production, film study, experimental film, critics and directors, sources of funds for filmmaking, local resources, periodicals for filmmaking classes, sources of free films, and…

  10. Dosimetric characterization of a novel intracavitary mold applicator for 192Ir high dose rate endorectal brachytherapy treatment.

    PubMed

    Poon, Emily; Reniers, Brigitte; Devic, Slobodan; Vuong, Té; Verhaegen, Frank

    2006-12-01

    The dosimetric properties of a novel intracavitary mold applicator for 192Ir high dose rate (HDR) endorectal cancer treatment have been investigated using Monte Carlo (MC) simulations and experimental methods. The 28 cm long applicator has a flexible structure made of silicone rubber for easy passage into cavities with deep-seated tumors. It consists of eight source catheters arranged around a central cavity for shielding insertion, and is compatible for use with an endocavitary balloon. A phase space model of the HDR source has been validated for dose calculations using the GEANT4 MC code. GAFCHROMIC EBT model film was used to measure dose distributions in water around shielded and unshielded applicators with two loading configurations, and to quantify the shielding effect of a balloon injected with an iodine solution (300 mg I/mL). The film calibration procedure was performed in water using an 192Ir HDR source. Ionization chamber measurements in a Lucite phantom show that placing a tungsten rod in the applicator attenuates the dose in the shielded region by up to 85%. Inserting the shielded applicator into a water-filled balloon pushes the neighboring tissues away from the radiation source, and the resulting geometric displacement reduces the dose by up to 53%; another 8% dose reduction can be achieved when the balloon is injected with an iodine solution. All experimental results agree with the GEANT4 calculations within measurement uncertainties. PMID:17278803

  11. Dosimetric characterization of a novel intracavitary mold applicator for {sup 192}Ir high dose rate endorectal brachytherapy treatment

    SciTech Connect

    Poon, Emily; Reniers, Brigitte; Devic, Slobodan; Vuong, Te; Verhaegen, Frank

    2006-12-15

    The dosimetric properties of a novel intracavitary mold applicator for {sup 192}Ir high dose rate (HDR) endorectal cancer treatment have been investigated using Monte Carlo (MC) simulations and experimental methods. The 28 cm long applicator has a flexible structure made of silicone rubber for easy passage into cavities with deep-seated tumors. It consists of eight source catheters arranged around a central cavity for shielding insertion, and is compatible for use with an endocavitary balloon. A phase space model of the HDR source has been validated for dose calculations using the GEANT4 MC code. GAFCHROMIC trade mark sign EBT model film was used to measure dose distributions in water around shielded and unshielded applicators with two loading configurations, and to quantify the shielding effect of a balloon injected with an iodine solution (300 mg I/mL). The film calibration procedure was performed in water using an {sup 192}Ir HDR source. Ionization chamber measurements in a Lucite phantom show that placing a tungsten rod in the applicator attenuates the dose in the shielded region by up to 85%. Inserting the shielded applicator into a water-filled balloon pushes the neighboring tissues away from the radiation source, and the resulting geometric displacement reduces the dose by up to 53%; another 8% dose reduction can be achieved when the balloon is injected with an iodine solution. All experimental results agree with the GEANT4 calculations within measurement uncertainties.

  12. A novel technique for measuring the low-dose envelope of pencil-beam scanning spot profiles

    NASA Astrophysics Data System (ADS)

    Lin, Liyong; Ainsley, Christopher G.; Mertens, Thierry; De Wilde, Olivier; Talla, Patrick T.; McDonough, James E.

    2013-06-01

    To investigate the profile measurement capabilities of an IBA-Dosimetry scintillation detector and to assess its feasibility for determining the low-intensity tails of pencil-beam scanning spots, the responses of the scintillation detector and Gafchromic EBT2 film to a 115 MeV proton spot were measured in-air at the isocenter. Pairs of irradiations were made: one lower-level irradiation insufficient to cause saturation, and one higher-level irradiation which deliberately saturated the central region of the spot, but provided magnification of the tails. By employing the pair/magnification technique, agreement between the film and scintillation detector measurements of the spot profile can be extended from 4% of the central spot dose down to 0.01%. Gamma analysis between these measurements shows 95% and 99% agreement within a ±9 cm bound using criteria of 3 mm/3% and 5 mm/5%, respectively. Above 4%, our 115 MeV proton spot can be well-described by Gaussian function; below 4%, non-Gaussian, diamond-shaped tails predominate.

  13. On the Use of Optically Stimulated Luminescent Dosimeter for Surface Dose Measurement during Radiotherapy.

    PubMed

    Yusof, Fasihah Hanum; Ung, Ngie Min; Wong, Jeannie Hsiu Ding; Jong, Wei Loong; Ath, Vannyat; Phua, Vincent Chee Ee; Heng, Siew Ping; Ng, Kwan Hoong

    2015-01-01

    This study was carried out to investigate the suitability of using the optically stimulated luminescence dosimeter (OSLD) in measuring surface dose during radiotherapy. The water equivalent depth (WED) of the OSLD was first determined by comparing the surface dose measured using the OSLD with the percentage depth dose at the buildup region measured using a Markus ionization chamber. Surface doses were measured on a solid water phantom using the OSLD and compared against the Markus ionization chamber and Gafchromic EBT3 film measurements. The effect of incident beam angles on surface dose was also studied. The OSLD was subsequently used to measure surface dose during tangential breast radiotherapy treatments in a phantom study and in the clinical measurement of 10 patients. Surface dose to the treated breast or chest wall, and on the contralateral breast were measured. The WED of the OSLD was found to be at 0.4 mm. For surface dose measurement on a solid water phantom, the Markus ionization chamber measured 15.95% for 6 MV photon beam and 12.64% for 10 MV photon beam followed by EBT3 film (23.79% and 17.14%) and OSLD (37.77% and 25.38%). Surface dose increased with the increase of the incident beam angle. For phantom and patient breast surface dose measurement, the response of the OSLD was higher than EBT3 film. The in-vivo measurements were also compared with the treatment planning system predicted dose. The OSLD measured higher dose values compared to dose at the surface (Hp(0.0)) by a factor of 2.37 for 6 MV and 2.01 for 10 MV photon beams, respectively. The measurement of absorbed dose at the skin depth of 0.4 mm by the OSLD can still be a useful tool to assess radiation effects on the skin dermis layer. This knowledge can be used to prevent and manage potential acute skin reaction and late skin toxicity from radiotherapy treatments. PMID:26052690

  14. On the Use of Optically Stimulated Luminescent Dosimeter for Surface Dose Measurement during Radiotherapy

    PubMed Central

    Yusof, Fasihah Hanum; Ung, Ngie Min; Wong, Jeannie Hsiu Ding; Jong, Wei Loong; Ath, Vannyat; Phua, Vincent Chee Ee; Heng, Siew Ping; Ng, Kwan Hoong

    2015-01-01

    This study was carried out to investigate the suitability of using the optically stimulated luminescence dosimeter (OSLD) in measuring surface dose during radiotherapy. The water equivalent depth (WED) of the OSLD was first determined by comparing the surface dose measured using the OSLD with the percentage depth dose at the buildup region measured using a Markus ionization chamber. Surface doses were measured on a solid water phantom using the OSLD and compared against the Markus ionization chamber and Gafchromic EBT3 film measurements. The effect of incident beam angles on surface dose was also studied. The OSLD was subsequently used to measure surface dose during tangential breast radiotherapy treatments in a phantom study and in the clinical measurement of 10 patients. Surface dose to the treated breast or chest wall, and on the contralateral breast were measured. The WED of the OSLD was found to be at 0.4 mm. For surface dose measurement on a solid water phantom, the Markus ionization chamber measured 15.95% for 6 MV photon beam and 12.64% for 10 MV photon beam followed by EBT3 film (23.79% and 17.14%) and OSLD (37.77% and 25.38%). Surface dose increased with the increase of the incident beam angle. For phantom and patient breast surface dose measurement, the response of the OSLD was higher than EBT3 film. The in-vivo measurements were also compared with the treatment planning system predicted dose. The OSLD measured higher dose values compared to dose at the surface (Hp(0.0)) by a factor of 2.37 for 6 MV and 2.01 for 10 MV photon beams, respectively. The measurement of absorbed dose at the skin depth of 0.4 mm by the OSLD can still be a useful tool to assess radiation effects on the skin dermis layer. This knowledge can be used to prevent and manage potential acute skin reaction and late skin toxicity from radiotherapy treatments. PMID:26052690

  15. Three-dimensional dosimetry of small megavoltage radiation fields using radiochromic gels and optical CT scanning

    NASA Astrophysics Data System (ADS)

    Babic, Steven; McNiven, Andrea; Battista, Jerry; Jordan, Kevin

    2009-04-01

    The dosimetry of small fields as used in stereotactic radiotherapy, radiosurgery and intensity-modulated radiation therapy can be challenging and inaccurate due to partial volume averaging effects and possible disruption of charged particle equilibrium. Consequently, there exists a need for an integrating, tissue equivalent dosimeter with high spatial resolution to avoid perturbing the radiation beam and artificially broadening the measured beam penumbra. In this work, radiochromic ferrous xylenol-orange (FX) and leuco crystal violet (LCV) micelle gels were used to measure relative dose factors (RDFs), percent depth dose profiles and relative lateral beam profiles of 6 MV x-ray pencil beams of diameter 28.1, 9.8 and 4.9 mm. The pencil beams were produced via stereotactic collimators mounted on a Varian 2100 EX linear accelerator. The gels were read using optical computed tomography (CT). Data sets were compared quantitatively with dosimetric measurements made with radiographic (Kodak EDR2) and radiochromic (GAFChromic® EBT) film, respectively. Using a fast cone-beam optical CT scanner (Vista™), corrections for diffusion in the FX gel data yielded RDFs that were comparable to those obtained by minimally diffusing LCV gels. Considering EBT film-measured RDF data as reference, cone-beam CT-scanned LCV gel data, corrected for scattered stray light, were found to be in agreement within 0.5% and -0.6% for the 9.8 and 4.9 mm diameter fields, respectively. The validity of the scattered stray light correction was confirmed by general agreement with RDF data obtained from the same LCV gel read out with a laser CT scanner that is less prone to the acceptance of scattered stray light. Percent depth dose profiles and lateral beam profiles were found to agree within experimental error for the FX gel (corrected for diffusion), LCV gel (corrected for scattered stray light), and EBT and EDR2 films. The results from this study reveal that a three-dimensional dosimetry method

  16. Evaporated VOx Thin Films

    NASA Astrophysics Data System (ADS)

    Stapinski, Tomasz; Leja, E.

    1989-03-01

    VOx thin films on glass were obtained by thermal evaporation of V205, powder. The structural investigations were carried out with the use of X-ray diffractometer. The electrical properties of the film were examined by means of temperature measurements of resistivity for the samples heat-treated in various conditions. Optical transmission and reflection spectra of VOX films of various composition showed the influence of the heat treatment.

  17. Clinical careers film.

    PubMed

    2015-09-01

    Those interested in developing clinical academic careers might be interested in a short animated film by Health Education England (HEE) and the National Institute for Health Research. The three-minute film, a frame from which is shown below, describes the sort of opportunities that are on offer to all professionals as part of the HEE's clinical academic careers framework. You can view the film on YouTube at tinyurl.com/pelb95c. PMID:26309005

  18. Renaissance of the Film.

    ERIC Educational Resources Information Center

    Bellone, Julius, Ed.

    The post-World War II period was one of the liveliest in the history of the cinema. This is a collection of 33 critical articles on some of the best films of the perd. Most of the essays explicate the themes and symbols of the films. The essays deal with these films: "The Apu Trilogy,""L'Avventura,""Balthazar,""Blow-Up,""Bonnie and Clyde," Citizen…

  19. Protective overcoating of films

    NASA Technical Reports Server (NTRS)

    Maas, K. A.

    1972-01-01

    Kodak Film Type SO-212 was emulsion overcoated with gelatin and lacquer to evaluate the feasibility of application of the coatings, any image degradation, and the relative protection offered against abrasion. Evaluated were: Eastman motion picture film lacquer Type 485, water solutions of Eastman purified Calfskin gelatin, and experimental Eastman gelatin stripping film of 4 and 6 microns. Conclusions reached were: (1) All coatings can be applied with relative ease with the only limitation being that of equipment. (2) None of the coatings degrade the processed image. (3) All of the coatings provide protection to the emulsion. These conclusions apply to any film which may be considered for overcoating.

  20. Nanocarbonic transparent conductive films.

    PubMed

    Roth, Siegmar; Park, Hye Jin

    2010-07-01

    This tutorial review discusses the contradictory material properties of electrical conductivity and optical transparency for the examples of graphene films and carbon nanotube networks. It is argued that for homogeneous films both properties are linked by basic laws of physics and that for perfect monoatomic layers conductivity and transparency can be calculated from the fine structure constant. To beat these limitations, inhomogeneous films are required, such as graphene with an array of holes or nanotube networks. An overview is given on literature values of transparency and conductivity, both for graphene films and for nanotube networks. PMID:20502813

  1. Historical film processing

    NASA Astrophysics Data System (ADS)

    Wu, Yi; Suter, David

    1995-08-01

    This paper describes work using historical film material, including what is believed to be the world's first feature length film. The digital processing of historical film material permits many new facilities: digital restoration, electronic storage, automated indexing, and electronic delivery to name a few. Although the work aims ultimately to support all of the previously mentioned facilities, this paper concentrated upon automatic scene change detection, brightness correction, and frame registration. These processes are fundamental to a more complete and complex processing system, but, by themselves, could be immediately used in computer-assisted film cataloging.

  2. The Possibility of Film Criticism.

    ERIC Educational Resources Information Center

    Poague, Leland; Cadbury, William

    1989-01-01

    Examines the role of critical language in film criticism. Compares and contrasts Monroe Beardsley's philosophy on film aesthetics with the New Criticism. Outlines some of the contributions Beardsley has made to the study of film criticism. (KM)

  3. Film-Makers' Cooperative Catalogue Number Six.

    ERIC Educational Resources Information Center

    Film-Makers' Cooperative, New York, NY.

    The Film-Makers' Cooperative services as a film depository and rental agency for any film-makers who may choose to make their work available to the general public. Approximately 1,600 films are listed in film-maker sequence. Excerpts of film reviews are provided for each entry. An index by film title and policies and procedures for film rental are…

  4. Microfilm--Which Film Type, Which Application?

    ERIC Educational Resources Information Center

    Dodson, Suzanne Cates

    1985-01-01

    Report on characteristics of different kinds of microfilm available indicates proper film for specific needs. Silver halide and nonsilver films, diazo film, vesicular film, reaction of films to light, effect of heat and humidity on films, film susceptibility to scratching, and potential longevity of film types are covered. (35 references) (EJS)

  5. SU-E-J-101: Retroactive Calculation of TLD and Film Dose in Anthropomorphic Phantom as Assessment of Updated TPS Performance

    SciTech Connect

    Alkhatib, H; Oves, S

    2014-06-01

    Purpose: To demonstrate a quick and comprehensive method verifying the accuracy of the updated dose model by recalculating dose distribution in an anthropomorphic phantom with a new version of the TPS and comparing the results to measured values. Methods: CT images and IMRT plan of an RPC anthropomorphic head phantom, previously calculated by Pinnacle 9.0, was re-computed using Pinnacle 9.2 and 9.6. The dosimeters within the phantom include four TLD capsules representing a primary PTV, two TLD capsules representing a secondary PTV, and two TLD capsules representing an organ at risk. Also included were three sheets of Gafchromic film. Performance of the updated TPS version was assessed by recalculating point doses and dose profiles corresponding to TLD and film position respectively and then comparing the results to reported values by the RPC. Results: Comparing calculated doses to reported measured doses from the RPC yielded an average disagreement of 1.48%, 2.04% and 2.10% for versions 9.0, 9.2, 9.6 respectively. Computed doses points all meet the RPC's passing criteria with the exception of the point representing the superior organ at risk in version 9.6. However, qualitative analysis of the recalculated dose profiles showed improved agreement with those of the RPC, especially in the penumbra region. Conclusion: This work has demonstrated the calculation results of Pinnacle 9.2 and 9.6 vs 9.0 version. Additionally, this study illustrates a method for the user to gain confidence upgrade to a newer version of the treatment planning system.

  6. SU-E-T-160: Evaluation of Accuracy for Target Margin Size Obtained From CBCT On Lung SBRT Based On Film Dosimetry

    SciTech Connect

    Lee, S; Lee, M; Kim, M; Suh, T; Park, J; Park, S

    2014-06-01

    Purpose: To analysis delivered dose on target using gafchromic films for evaluating accuracy of target margin size obtained from cone beam computed tomography (CBCT) during lung stereotactic body radiation therapy (SBRT) Methods: The phantom consists of measurement part and driving part. The motor of Quasar motion phantom (Modus Medical Devices Inc, London, ON, Canada) was used for driving part and we developed measurement part which consist of cork cylindrical body and acrylic target with radiochromic film inserted into central and both ends of acrylic target. In this study lung SBRT cases through both four dimensional computed tomography (4DCT) and CBCT were selected. Target contouring including margin based on 4DCT is defined with a 1 cm margin around gross tumor volume (GTV) in all directions except for inferior direction. The moving range in inferior direction was larger than other directions thus, including 2 cm margin. In case of CBCT, the margin means blurring of target on CBCT images. This study was compared margin size determined through 4DCT and that of based on CBCT and we also evaluated dose profile and the length of margin in superior-inferior direction on CBCT compared with 4DCT. Results: The length of target including margin was 2.48 cm (based on CBCT) and 2.66 cm (based on 4DCT), respectively in superior-inferior direction. The difference of delivered dose on target between two margins was only within 1%. Conclusions: This study has shown the feasibility of determining target margin using CBCT for delivering more accurate prescription dose to lung cancer.

  7. Application of a dummy eye shield for electron treatment planning

    PubMed Central

    Kang, Sei-Kwon; Park, Soah; Hwang, Taejin; Cheong, Kwang-Ho; Han, Taejin; Kim, Haeyoung; Lee, Me-Yeon; Kim, Kyoung Ju; Oh, Do Hoon; Bae, Hoonsik

    2013-01-01

    Metallic eye shields have been widely used for near-eye treatments to protect critical regions, but have never been incorporated into treatment plans because of the unwanted appearance of the metal artifacts on CT images. The purpose of this work was to test the use of an acrylic dummy eye shield as a substitute for a metallic eye shield during CT scans. An acrylic dummy shield of the same size as the tungsten eye shield was machined and CT scanned. The BEAMnrc and the DOSXYZnrc were used for the Monte Carlo (MC) simulation, with the appropriate material information and density for the aluminum cover, steel knob and tungsten body of the eye shield. The Pinnacle adopting the Hogstrom electron pencil-beam algorithm was used for the one-port 6-MeV beam plan after delineation and density override of the metallic parts. The results were confirmed with the metal oxide semiconductor field effect transistor (MOSFET) detectors and the Gafchromic EBT2 film measurements. For both the maximum eyelid dose over the shield and the maximum dose under the shield, the MC results agreed with the EBT2 measurements within 1.7%. For the Pinnacle plan, the maximum dose under the shield agreed with the MC within 0.3%; however, the eyelid dose differed by –19.3%. The adoption of the acrylic dummy eye shield was successful for the treatment plan. However, the Pinnacle pencil-beam algorithm was not sufficient to predict the eyelid dose on the tungsten shield, and more accurate algorithms like MC should be considered for a treatment plan. PMID:22915776

  8. Application of a dummy eye shield for electron treatment planning.

    PubMed

    Kang, Sei-Kwon; Park, Soah; Hwang, Taejin; Cheong, Kwang-Ho; Han, Taejin; Kim, Haeyoung; Lee, Me-Yeon; Kim, Kyoung Ju; Oh, Do Hoon; Bae, Hoonsik

    2013-01-01

    Metallic eye shields have been widely used for near-eye treatments to protect critical regions, but have never been incorporated into treatment plans because of the unwanted appearance of the metal artifacts on CT images. The purpose of this work was to test the use of an acrylic dummy eye shield as a substitute for a metallic eye shield during CT scans. An acrylic dummy shield of the same size as the tungsten eye shield was machined and CT scanned. The BEAMnrc and the DOSXYZnrc were used for the Monte Carlo (MC) simulation, with the appropriate material information and density for the aluminum cover, steel knob and tungsten body of the eye shield. The Pinnacle adopting the Hogstrom electron pencil-beam algorithm was used for the one-port 6-MeV beam plan after delineation and density override of the metallic parts. The results were confirmed with the metal oxide semiconductor field effect transistor (MOSFET) detectors and the Gafchromic EBT2 film measurements. For both the maximum eyelid dose over the shield and the maximum dose under the shield, the MC results agreed with the EBT2 measurements within 1.7%. For the Pinnacle plan, the maximum dose under the shield agreed with the MC within 0.3%; however, the eyelid dose differed by -19.3%. The adoption of the acrylic dummy eye shield was successful for the treatment plan. However, the Pinnacle pencil-beam algorithm was not sufficient to predict the eyelid dose on the tungsten shield, and more accurate algorithms like MC should be considered for a treatment plan. PMID:22915776

  9. Water depth penetration film test

    NASA Technical Reports Server (NTRS)

    Lockwood, H. E.; Perry, L.; Sauer, G. E.; Lamar, N. T.

    1974-01-01

    As part of the National Aeronautics and Space Administration Earth Resources Program, a comparative and controlled evaluation of nine film-filter combinations was completed to establish the relative effectiveness in recording water subsurface detail if exposed from an aerial platform over a typical water body. The films tested, with one exception, were those which prior was suggested had potential. These included an experimental 2-layer positive color film, a 2-layer (minus blue layer) film, a normal 3-layer color film, a panchromatic black-and-white film, and a black-and-white infrared film. Selective filtration was used with all films.

  10. Thin-film forces in pseudoemulsion films

    SciTech Connect

    Bergeron, V.; Radke, C.J. |

    1991-06-01

    Use of foam for enhanced oil recovery (EOR) has shown recent success in steam-flooding field applications. Foam can also provide an effective barrier against gas coning in thin oil zones. Both of these applications stem from the unique mobility-control properties a stable foam possesses when it exists in porous media. Unfortunately, oil has a major destabilizing effect on foam. Therefore, it is important for EOR applications to understand how oil destroys foam. Studies all indicate that stabilization of the pseudoemulsion film is critical to maintain foam stability in the presence of oil. Hence, to aid in design of surfactant formulations for foam insensitivity to oil the authors pursue direct measurement of the thin-film or disjoining forces that stabilize pseudoemulsion films. Experimental procedures and preliminary results are described.

  11. Thick film hydrogen sensor

    DOEpatents

    Hoffheins, Barbara S.; Lauf, Robert J.

    1995-01-01

    A thick film hydrogen sensor element includes an essentially inert, electrically-insulating substrate having deposited thereon a thick film metallization forming at least two resistors. The metallization is a sintered composition of Pd and a sinterable binder such as glass frit. An essentially inert, electrically insulating, hydrogen impermeable passivation layer covers at least one of the resistors.

  12. Thick film hydrogen sensor

    DOEpatents

    Hoffheins, B.S.; Lauf, R.J.

    1995-09-19

    A thick film hydrogen sensor element includes an essentially inert, electrically-insulating substrate having deposited thereon a thick film metallization forming at least two resistors. The metallization is a sintered composition of Pd and a sinterable binder such as glass frit. An essentially inert, electrically insulating, hydrogen impermeable passivation layer covers at least one of the resistors. 8 figs.

  13. Authors on Film.

    ERIC Educational Resources Information Center

    Geduld, Harry M., Ed.

    Different authors' attitudes toward film are revealed through five different sections of this book: (1) articles, essays, and reviews pertaining to the silent cinema and the transition to sound; (2) general statements on the film medium or filmmakers and their messages; (3) essays dealing with the problems, involvements, and reflections of the…

  14. Construction of Meaning: Film.

    ERIC Educational Resources Information Center

    Pryluck, Calvin

    1995-01-01

    Notes that film has no clear set of rules, unlike all languages, which are deductive systems interpreted according to clear sets of rules. Suggests that film is an inductive system whose interpretation is based on a general understanding of events depicted as modified by production variables such as lighting, camera angles, and the context of…

  15. Stabilized chromium oxide film

    DOEpatents

    Nyaiesh, A.R.; Garwin, E.L.

    1986-08-04

    Stabilized air-oxidized chromium films deposited on high-power klystron ceramic windows and sleeves having a thickness between 20 and 150A are useful in lowering secondary electron emission yield and in avoiding multipactoring and window failure due to overheating. The ceramic substrate for the film is chosen from alumina, sapphire or beryllium oxide.

  16. Film Study Hang Ups.

    ERIC Educational Resources Information Center

    Grenier, Charles F.

    1969-01-01

    The interest and delight which students find in film should be preserved from a teacher's excessive zeal to analyze and explain. As the beauty of poetry is frequently diminished through exhaustive analyses of similes, rhyme schemes, and other technical devices, the value of film to high school students can be weakened through too great an emphasis…

  17. GPN Film Catalog 1972.

    ERIC Educational Resources Information Center

    Nebraska Univ., Lincoln. Great Plains National Instructional Television Library.

    The films described in this catalog were produced by schools or school-related organizations and were designed to meet the "relevant needs expressed by a broad spectrum of media personnel, students, and educators across the country." The catalog describes seventeen series and eight single films. For each of the series a description is presented…

  18. Elements of Film.

    ERIC Educational Resources Information Center

    Bobker, Lee R.

    A film is the successful combination of two distinct groups of elements: (1) the technical elements by which the film is made (camera, lighting, sound and editing) and (2) the esthetic elements that transform the craft into an art. This book attempts to combine the study of these elements by providing technical information about the process of…

  19. Creative Film-Making.

    ERIC Educational Resources Information Center

    Smallman, Kirk

    The fundamentals of motion picture photography are introduced with a physiological explanation for the illusion of motion in a film. Film stock formats and emulsions, camera features, and lights are listed and described. Various techniques of exposure control are illustrated in terms of their effects. Photographing action with a stationary or a…

  20. Film and the Humanities.

    ERIC Educational Resources Information Center

    O'Connor, John E., Ed.

    This report on a conference, which brought together representatives of various humanistic disciplines to explore the cross-disciplinary appeal of film study as well as the use of film in stimulating scholarship and teaching, includes a narrative summary of the day's conversations and issues raised, as well as of reprints of articles that suggest…

  1. Filming for Television.

    ERIC Educational Resources Information Center

    Englander, A. Arthur; Petzold, Paul

    Film makers, professional or amateur, will find in this volume an extensive discussion of the adaptation of film technique to television work, of the art of the camera operator, and of the productive relationships between people, organization, and hardware. Chapters include "The Beginnings," an overview of the interrelationship between roles in…

  2. Stabilized chromium oxide film

    DOEpatents

    Garwin, Edward L.; Nyaiesh, Ali R.

    1988-01-01

    Stabilized air-oxidized chromium films deposited on high-power klystron ceramic windows and sleeves having a thickness between 20 and 150.ANG. are useful in lowering secondary electron emission yield and in avoiding multipactoring and window failure due to overheating. The ceramic substrate for the film is chosen from alumina, sapphire or beryllium oxide.

  3. On Teaching Ethnographic Film

    ERIC Educational Resources Information Center

    Clarfield, Geoffrey

    2013-01-01

    The author of this article, a developmental anthropologist, illustrates how the instructor can use ethnographic films to enhance the study of anthropology and override notions about the scope and efficacy of Western intervention in the Third World, provided the instructor places such films in their proper historical and cultural context. He…

  4. Film Canister Science

    ERIC Educational Resources Information Center

    Ferstl, Andrew; Schneider, Jamie L.

    2007-01-01

    Opaque film canisters are readily available, cheap, and useful for scientific inquiry in the classroom. They can also be surprisingly versatile and useful as a tool for stimulating scientific inquiry. In this article, the authors describe inquiry activities using film canisters for preservice teachers, including a "black box" activity and several…

  5. Protolytic carbon film technology

    SciTech Connect

    Renschler, C.L.; White, C.A.

    1996-04-01

    This paper presents a technique for the deposition of polyacrylonitrile (PAN) on virtually any surface allowing carbon film formation with only the caveat that the substrate must withstand carbonization temperatures of at least 600 degrees centigrade. The influence of processing conditions upon the structure and properties of the carbonized film is discussed. Electrical conductivity, microstructure, and morphology control are also described.

  6. Critical Approaches to Film.

    ERIC Educational Resources Information Center

    Bywater, Timothy Robert

    This study deals primarily with recent academically oriented critical material, but it also embraces the range of film criticism that has been written for the mass audience in newspapers and periodicals. The study considers eight types of critical approaches to analyzing film: the journalistic approach, which contains both a reportorial-review and…

  7. Determining Film Art.

    ERIC Educational Resources Information Center

    Bullis, Roger

    1974-01-01

    The criteria by which films can and should be analyzed as art are discussed in this paper. A triangular model of theme-form-content is presented with form given greater significance than is usually the case in film criticism. The form-content-theme synthesis is the process in which theme is made clear by means of form and content within an…

  8. Films and Feelings.

    ERIC Educational Resources Information Center

    Gaffney, Maureen, Ed.

    1981-01-01

    Children's emotional responses to films are the focus of the four articles in this issue designed for media specialists and educators. Following an editorial discussing the responsibilities of media as put forth by Bruno Bettleheim, the first article presents a methodology and rationale for using story films to encourage children's exploration of…

  9. Discovery in Film.

    ERIC Educational Resources Information Center

    Heyer, Robert; Meyer, Anthony

    Seventy-eight "classic" short, non-feature films, cataloged according to six themes--communication, freedom, love, peace, happiness, and the underground--are discussed in this guide. The standard critique for each film has a fourfold purpose: (1) to indicate content and style; (2) to pose questions which provoke discussion; (3) to suggest resource…

  10. FAA Film Catalog.

    ERIC Educational Resources Information Center

    Federal Aviation Administration (DOT), Washington, DC.

    Some 75 films from the U.S. Department of Transportation's Federal Aviation Administration are listed in this catalog. Topics dealt with include aerodynamics, airports, aviation history and careers, flying clubs, navigation and weather. Most of the films are 16mm sound and color productions. Filmstrips requiring a 35mm projector and phonograph or…

  11. Dental Training Films.

    ERIC Educational Resources Information Center

    Veterans Administration Medical Center, Washington, DC.

    This dental training films catalog is organized into two sections. Section I is a category listing of the films by number and title, indexed according to generalized headings; categories are as follow: anatomy, articulator systems, complete dentures, dental assisting, dental laboratory technology, dental materials, dental office emergencies,…

  12. Film Making in Schools.

    ERIC Educational Resources Information Center

    Lowndes, Douglas

    A program which used practical film study to extend powers of observation and comment and to help young people (ages 12 through 16) develop an understanding of contemporary society is described in this manual. The role of film in school curricula and its integration with other studies is discussed in an opening section. The next section contains a…

  13. Abstract Film and Beyond.

    ERIC Educational Resources Information Center

    Le Grice, Malcolm

    A theoretical and historical account of the main preoccupations of makers of abstract films is presented in this book. The book's scope includes discussion of nonrepresentational forms as well as examination of experiments in the manipulation of time in films. The ten chapters discuss the following topics: art and cinematography, the first…

  14. Bursting of soap films

    NASA Astrophysics Data System (ADS)

    Chowdhury, Debashish; Stauffer, Dietrich

    1992-07-01

    Soap films consist of thin films of water in between two monolayers of amphiphilic molecules. Newton black films (NBFs) are the thinnest possible soap films. We have developed a microscopic model of NBFs; this model is a variant of the Widom model for microemulsions. By carrying out Monte Carlo simulations of this model, we have investigated the dependence of the lifetime of the NBFs on (a) the initial concentration of the amphiphilic molecules, (b) the temperature and (c) the bending rigidity of the constituent amphiphilic monolayers. We compare our results with the corresponding experimental observations and suggest further specific experiments. We establish that the “edge energy” of the model bilayer tends to stabilize the NBF; a similar mechanism leads to the well known phenomenon of “self-healing” of small enough holes in pierced vesicles. We also review the laws of growth of holes in soap films during rupture. Finally, we speculate on some other possible applications of our ideas.

  15. Selective inorganic thin films

    SciTech Connect

    Phillips, M.L.F.; Weisenbach, L.A.; Anderson, M.T.

    1995-05-01

    This project is developing inorganic thin films as membranes for gas separation applications, and as discriminating coatings for liquid-phase chemical sensors. Our goal is to synthesize these coatings with tailored porosity and surface chemistry on porous substrates and on acoustic and optical sensors. Molecular sieve films offer the possibility of performing separations involving hydrogen, air, and natural gas constituents at elevated temperatures with very high separation factors. We are focusing on improving permeability and molecular sieve properties of crystalline zeolitic membranes made by hydrothermally reacting layered multicomponent sol-gel films deposited on mesoporous substrates. We also used acoustic plate mode (APM) oscillator and surface plasmon resonance (SPR) sensor elements as substrates for sol-gel films, and have both used these modified sensors to determine physical properties of the films and have determined the sensitivity and selectivity of these sensors to aqueous chemical species.

  16. The National Film Registry: Acquiring Our Film Heritage.

    ERIC Educational Resources Information Center

    Ziegler, Roy A.

    The National Film Registry, which is primarily a designated list of films to be preserved by the Library of Congress, is also a valuable tool for selecting "films that are culturally, historically, and aesthetically significant." Following a brief discussion of the history and selection process of the National Film Registry, Southeast Missouri…

  17. Quantitative assessment of the accuracy of dose calculation using pencil beam and Monte Carlo algorithms and requirements for clinical quality assurance

    SciTech Connect

    Ali, Imad; Ahmad, Salahuddin

    2013-10-01

    To compare the doses calculated using the BrainLAB pencil beam (PB) and Monte Carlo (MC) algorithms for tumors located in various sites including the lung and evaluate quality assurance procedures required for the verification of the accuracy of dose calculation. The dose-calculation accuracy of PB and MC was also assessed quantitatively with measurement using ionization chamber and Gafchromic films placed in solid water and heterogeneous phantoms. The dose was calculated using PB convolution and MC algorithms in the iPlan treatment planning system from BrainLAB. The dose calculation was performed on the patient's computed tomography images with lesions in various treatment sites including 5 lungs, 5 prostates, 4 brains, 2 head and necks, and 2 paraspinal tissues. A combination of conventional, conformal, and intensity-modulated radiation therapy plans was used in dose calculation. The leaf sequence from intensity-modulated radiation therapy plans or beam shapes from conformal plans and monitor units and other planning parameters calculated by the PB were identical for calculating dose with MC. Heterogeneity correction was considered in both PB and MC dose calculations. Dose-volume parameters such as V95 (volume covered by 95% of prescription dose), dose distributions, and gamma analysis were used to evaluate the calculated dose by PB and MC. The measured doses by ionization chamber and EBT GAFCHROMIC film in solid water and heterogeneous phantoms were used to quantitatively asses the accuracy of dose calculated by PB and MC. The dose-volume histograms and dose distributions calculated by PB and MC in the brain, prostate, paraspinal, and head and neck were in good agreement with one another (within 5%) and provided acceptable planning target volume coverage. However, dose distributions of the patients with lung cancer had large discrepancies. For a plan optimized with PB, the dose coverage was shown as clinically acceptable, whereas in reality, the MC showed a

  18. Chiral atomically thin films

    NASA Astrophysics Data System (ADS)

    Kim, Cheol-Joo; Sánchez-Castillo, A.; Ziegler, Zack; Ogawa, Yui; Noguez, Cecilia; Park, Jiwoong

    2016-06-01

    Chiral materials possess left- and right-handed counterparts linked by mirror symmetry. These materials are useful for advanced applications in polarization optics, stereochemistry and spintronics. In particular, the realization of spatially uniform chiral films with atomic-scale control of their handedness could provide a powerful means for developing nanodevices with novel chiral properties. However, previous approaches based on natural or grown films, or arrays of fabricated building blocks, could not offer a direct means to program intrinsic chiral properties of the film on the atomic scale. Here, we report a chiral stacking approach, where two-dimensional materials are positioned layer-by-layer with precise control of the interlayer rotation (θ) and polarity, resulting in tunable chiral properties of the final stack. Using this method, we produce left- and right-handed bilayer graphene, that is, a two-atom-thick chiral film. The film displays one of the highest intrinsic ellipticity values (6.5 deg μm–1) ever reported, and a remarkably strong circular dichroism (CD) with the peak energy and sign tuned by θ and polarity. We show that these chiral properties originate from the large in-plane magnetic moment associated with the interlayer optical transition. Furthermore, we show that we can program the chiral properties of atomically thin films layer-by-layer by producing three-layer graphene films with structurally controlled CD spectra.

  19. Chiral atomically thin films.

    PubMed

    Kim, Cheol-Joo; Sánchez-Castillo, A; Ziegler, Zack; Ogawa, Yui; Noguez, Cecilia; Park, Jiwoong

    2016-06-01

    Chiral materials possess left- and right-handed counterparts linked by mirror symmetry. These materials are useful for advanced applications in polarization optics, stereochemistry and spintronics. In particular, the realization of spatially uniform chiral films with atomic-scale control of their handedness could provide a powerful means for developing nanodevices with novel chiral properties. However, previous approaches based on natural or grown films, or arrays of fabricated building blocks, could not offer a direct means to program intrinsic chiral properties of the film on the atomic scale. Here, we report a chiral stacking approach, where two-dimensional materials are positioned layer-by-layer with precise control of the interlayer rotation (θ) and polarity, resulting in tunable chiral properties of the final stack. Using this method, we produce left- and right-handed bilayer graphene, that is, a two-atom-thick chiral film. The film displays one of the highest intrinsic ellipticity values (6.5 deg μm(-1)) ever reported, and a remarkably strong circular dichroism (CD) with the peak energy and sign tuned by θ and polarity. We show that these chiral properties originate from the large in-plane magnetic moment associated with the interlayer optical transition. Furthermore, we show that we can program the chiral properties of atomically thin films layer-by-layer by producing three-layer graphene films with structurally controlled CD spectra. PMID:26900756

  20. Film Scriptwriting: A Practical Manual.

    ERIC Educational Resources Information Center

    Swain, Dwight V.

    Dealing with both documentary and feature films, this book is a guide to using particular tools and procedures in developing ideas and concepts for writing film scripts. Part one deals with the factual, or documentary, film and discusses the proposal outline, film treatment, sequence outline, shooting script, and narration writing. Part two…

  1. Using Films in Literature Courses.

    ERIC Educational Resources Information Center

    Littell, Katherine M.

    1976-01-01

    Films are becoming recognized as an important medium of foreign language instruction. Although there is a lack of instructional films in German, an abundance of excellent feature films is available through the German Embassy and various consulates. Well-edited scenarios are needed, however, in order to make using these films an effective teaching…

  2. Diagnosing Disputes in Film Criticism.

    ERIC Educational Resources Information Center

    Schultz, Robert A.

    1979-01-01

    By classifying and analyzing actual disputes in film criticism, the author considers the following questions: Are there connections between film theory and film criticism? If so, which are healthy and which are diseased? If not, what alternative healthy function might film theory have? (Author/SJL)

  3. The Art of the Film.

    ERIC Educational Resources Information Center

    Lindgren, Ernest

    The author prefaces his consideration of films as an art form with a discussion of the mechanics of filmmaking. He describes the division of talent on a movie set, details the history of the tools of filmmakers, and explains the production and reproduction of a film. The influence of film techniques on plot development in a fiction film is…

  4. Discovery in Film, Book Two.

    ERIC Educational Resources Information Center

    Gordon, Malcolm W.

    Approximately 80 16 millimeter (16mm) short films are reviewed in this introduction and guide which attempts to be comprehensive in touching the major areas and styles of 16mm films now being produced. An attempt is made to describe as carefully as possible the style and content of each film and suggest ways in which the films might be used. Films…

  5. Negative birefringent polyimide films

    NASA Technical Reports Server (NTRS)

    Harris, Frank W. (Inventor); Cheng, Stephen Z. D. (Inventor)

    1994-01-01

    A negative birefringent film, useful in liquid crystal displays, and a method for controlling the negative birefringence of a polyimide film is disclosed which allows the matching of an application to a targeted amount of birefringence by controlling the degree of in-plane orientation of the polyimide by the selection of functional groups within both the diamine and dianhydride segments of the polyimide which affect the polyimide backbone chain rigidity, linearity, and symmetry. The higher the rigidity, linearity and symmetry of the polyimide backbone, the larger the value of the negative birefringence of the polyimide film.

  6. Thin film hydrogen sensor

    DOEpatents

    Cheng, Y.T.; Poli, A.A.; Meltser, M.A.

    1999-03-23

    A thin film hydrogen sensor includes a substantially flat ceramic substrate with first and second planar sides and a first substrate end opposite a second substrate end; a thin film temperature responsive resistor on the first planar side of the substrate proximate to the first substrate end; a thin film hydrogen responsive metal resistor on the first planar side of the substrate proximate to the fist substrate end and proximate to the temperature responsive resistor; and a heater on the second planar side of the substrate proximate to the first end. 5 figs.

  7. Ultrahard carbon nanocomposite films

    SciTech Connect

    SIEGAL,MICHAEL P.; TALLANT,DAVID R.; PROVENCIO,PAULA P.; OVERMYER,DONALD L.; SIMPSON,REGINA L.; MARTINEZ-MIRANDA,L.J.

    2000-01-27

    Modest thermal annealing to 600 C of diamondlike amorphous-carbon (a-C) films grown at room temperature results in the formation of carbon nanocomposites with hardness similar to diamond. These nanocomposite films consist of nanometer-sized regions of high density a-C embedded in an a-C matrix with a reduced density of 5--10%. The authors report on the evolution of density and bonding topologies as a function of annealing temperature. Despite a decrease in density, film hardness actually increases {approximately} 15% due to the development of the nanocomposite structure.

  8. Biomimetic thin film synthesis

    SciTech Connect

    Graff, G.L.; Campbell, A.A.; Gordon, N.R.

    1995-05-01

    The purpose of this program is to develop a new process for forming thin film coatings and to demonstrate that the biomimetic thin film technology developed at PNL is useful for industrial applications. In the biomimetic process, mineral deposition from aqueous solution is controlled by organic functional groups attached to the underlying substrate surface. The coatings process is simple, benign, inexpensive, energy efficient, and particularly suited for temperature sensitive substrate materials (such as polymers). In addition, biomimetic thin films can be deposited uniformly on complex shaped and porous substrates providing a unique capability over more traditional line-of-sight methods.

  9. Thin film hydrogen sensor

    DOEpatents

    Cheng, Yang-Tse; Poli, Andrea A.; Meltser, Mark Alexander

    1999-01-01

    A thin film hydrogen sensor, includes: a substantially flat ceramic substrate with first and second planar sides and a first substrate end opposite a second substrate end; a thin film temperature responsive resistor on the first planar side of the substrate proximate to the first substrate end; a thin film hydrogen responsive metal resistor on the first planar side of the substrate proximate to the fist substrate end and proximate to the temperature responsive resistor; and a heater on the second planar side of the substrate proximate to the first end.

  10. Amorphous diamond films

    DOEpatents

    Falabella, S.

    1998-06-09

    Amorphous diamond films having a significant reduction in intrinsic stress are prepared by biasing a substrate to be coated and depositing carbon ions thereon under controlled temperature conditions. 1 fig.

  11. Magnetron sputtered boron films

    DOEpatents

    Makowiecki, Daniel M.; Jankowski, Alan F.

    1998-01-01

    A method is described for the production of thin boron and titanium/boron films by magnetron sputter deposition. The amorphous boron films contain no morphological growth features, unlike those found when thin films are prepared by various physical vapor deposition processes. Magnetron sputter deposition method requires the use of a high density crystalline boron sputter target which is prepared by hot isostatic pressing. Thin boron films prepared by this method are useful for producing hardened surfaces, surfacing machine tools, etc. and for ultra-thin band pass filters as well as the low Z element in low Z/high Z optical components, such as mirrors which enhance reflectivity from grazing to normal incidence.

  12. Magnetron sputtered boron films

    DOEpatents

    Makowiecki, D.M.; Jankowski, A.F.

    1998-06-16

    A method is described for the production of thin boron and titanium/boron films by magnetron sputter deposition. The amorphous boron films contain no morphological growth features, unlike those found when thin films are prepared by various physical vapor deposition processes. Magnetron sputter deposition method requires the use of a high density crystalline boron sputter target which is prepared by hot isostatic pressing. Thin boron films prepared by this method are useful for producing hardened surfaces, surfacing machine tools, etc. and for ultra-thin band pass filters as well as the low Z element in low Z/high Z optical components, such as mirrors which enhance reflectivity from grazing to normal incidence. 8 figs.

  13. Nanostructured thermoplastic polyimide films

    SciTech Connect

    Aglan, Heshmat

    2015-05-19

    Structured films containing multi-walled carbon nanotubes ("MWCNTs") have enhanced mechanical performance in terms of strength, fracture resistance, and creep recovery of polyimide ("PI") films. Preferably, the loadings of MWCNTs can be in the range of 0.1 wt % to 0.5 wt %. The strength of the new PI films dried at 60.degree. C. increased by 55% and 72% for 0.1 wt % MWCNT and 0.5 wt % MWCNT loadings, respectively, while the fracture resistance increased by 23% for the 0.1 wt % MWCNTs and then decreases at a loading of 0.5 wt % MWCNTs. The films can be advantageously be created by managing a corresponding shift in the annealing temperature at which the maximum strength occurs as the MWCNT loadings increase.

  14. Quantitative film radiography

    SciTech Connect

    Devine, G.; Dobie, D.; Fugina, J.; Hernandez, J.; Logan, C.; Mohr, P.; Moss, R.; Schumacher, B.; Updike, E.; Weirup, D.

    1991-02-26

    We have developed a system of quantitative radiography in order to produce quantitative images displaying homogeneity of parts. The materials that we characterize are synthetic composites and may contain important subtle density variations not discernible by examining a raw film x-radiograph. In order to quantitatively interpret film radiographs, it is necessary to digitize, interpret, and display the images. Our integrated system of quantitative radiography displays accurate, high-resolution pseudo-color images in units of density. We characterize approximately 10,000 parts per year in hundreds of different configurations and compositions with this system. This report discusses: the method; film processor monitoring and control; verifying film and processor performance; and correction of scatter effects.

  15. TAMPERPROOF FILM BADGE

    DOEpatents

    Kocher, L.F.

    1958-10-01

    A persornel dosimeter film badge made of plastic, with provision for a picture of the wearer and an internal slide containing photographic film that is sensitive to various radiations, is described. Four windows made of differing material selectively attenuate alpha, beta, gamma rays, and neutrons so as to distinguish the particular type of radiation the wearer was subjected to. In addition, a lead shield has the identification number of the wearer perforated thereon so as to identify the film after processing. An internal magnetically actuated latch securely locks the slide within the body, and may be withdrawn only upon the external application of two strong magnetic forces in order to insure that the wearer or other curious persons will not accidentally expose the film to visual light.

  16. Sprites on Film

    NASA Video Gallery

    Filmed at 10,000 frames per second by Japan's NHK television, movies like this of electromagnetic bursts called "sprites" will help scientists better understand how weather high in the atmosphere r...

  17. Magnetoresistance of Au films

    SciTech Connect

    Zhang, D. L.; Song, X. H.; Zhang, X; Zhang, Xiaoguang

    2014-01-01

    Measurement of the magnetoresistance (MR) of Au films as a function of temperature and film thickness reveals a strong dependence on grain size distribution and clear violation of the Kohler s rule. Using a model of random resistor network, we show that this result can be explained if the MR arises entirely from inhomogeneity due to grain boundary scattering and thermal activation of grain boundary atoms.

  18. Polymer film composite transducer

    DOEpatents

    Owen, Thomas E.

    2005-09-20

    A composite piezoelectric transducer, whose piezoeletric element is a "ribbon wound" film of piezolectric material. As the film is excited, it expands and contracts, which results in expansion and contraction of the diameter of the entire ribbon winding. This is accompanied by expansion and contraction of the thickness of the ribbon winding, such that the sound radiating plate may be placed on the side of the winding.

  19. Multifunctional thin film surface

    SciTech Connect

    Brozik, Susan M.; Harper, Jason C.; Polsky, Ronen; Wheeler, David R.; Arango, Dulce C.; Dirk, Shawn M.

    2015-10-13

    A thin film with multiple binding functionality can be prepared on an electrode surface via consecutive electroreduction of two or more aryl-onium salts with different functional groups. This versatile and simple method for forming multifunctional surfaces provides an effective means for immobilization of diverse molecules at close proximities. The multifunctional thin film has applications in bioelectronics, molecular electronics, clinical diagnostics, and chemical and biological sensing.

  20. Magnetoresistance of Au films

    DOE PAGESBeta

    Zhang, D. L.; Song, X. H.; Zhang, X.; Zhang, Xiaoguang

    2014-12-10

    Measurement of the magnetoresistance (MR) of Au films as a function of temperature and film thickness reveals a strong dependence on grain size distribution and clear violation of the Kohler s rule. Using a model of random resistor network, we show that this result can be explained if the MR arises entirely from inhomogeneity due to grain boundary scattering and thermal activation of grain boundary atoms.

  1. Thin film tritium dosimetry

    DOEpatents

    Moran, Paul R.

    1976-01-01

    The present invention provides a method for tritium dosimetry. A dosimeter comprising a thin film of a material having relatively sensitive RITAC-RITAP dosimetry properties is exposed to radiation from tritium, and after the dosimeter has been removed from the source of the radiation, the low energy electron dose deposited in the thin film is determined by radiation-induced, thermally-activated polarization dosimetry techniques.

  2. Dosimetry of a Small-Animal Irradiation Model using a 6 MV Linear Accelerator

    SciTech Connect

    Fitch, F. Moran; Martinez-Davalos, A.; Garcia-Garduno, O. A.

    2010-12-07

    A custom made rat-like phantom was used to measure dose distributions using a 6 MV linear accelerator. The phantom has air cavities that simulate the lungs and cylindrical inserts that simulate the backbone. The calculated dose distributions were obtained with the BrainScan v.5.31 TPS software. For the irradiation two cases were considered: (a) near the region where the phantom has two air cavities that simulate the lungs, and (b) with an entirely uniform phantom. The treatment plan consisted of two circular cone arcs that imparted a 500 cGy dose to a simulated lesion in the backbone. We measured dose distributions using EBT2 GafChromic film and an Epson Perfection V750 scanner working in transmission mode. Vertical and horizontal profiles, isodose curves from 50 to 450 cGy, dose and distance to agreement (DTA) histograms and Gamma index were obtained to compare the dose distributions using DoseLab v4.11. As a result, these calculations show very good agreement between calculated and measured dose distribution in both cases. With a 2% 2 mm criteria 100% of the points pass the Gamma test for the uniform case, while 98.9% of the points do it for the lungs case.

  3. Dosimetry of a Small-Animal Irradiation Model using a 6 MV Linear Accelerator

    NASA Astrophysics Data System (ADS)

    Fitch, F. Morán; García-Garduño, O. A.; Martínez-Dávalos, A.

    2010-12-01

    A custom made rat-like phantom was used to measure dose distributions using a 6 MV linear accelerator. The phantom has air cavities that simulate the lungs and cylindrical inserts that simulate the backbone. The calculated dose distributions were obtained with the BrainScan v.5.31 TPS software. For the irradiation two cases were considered: (a) near the region where the phantom has two air cavities that simulate the lungs, and (b) with an entirely uniform phantom. The treatment plan consisted of two circular cone arcs that imparted a 500 cGy dose to a simulated lesion in the backbone. We measured dose distributions using EBT2 GafChromic film and an Epson Perfection V750 scanner working in transmission mode. Vertical and horizontal profiles, isodose curves from 50 to 450 cGy, dose and distance to agreement (DTA) histograms and Gamma index were obtained to compare the dose distributions using DoseLab v4.11. As a result, these calculations show very good agreement between calculated and measured dose distribution in both cases. With a 2% 2 mm criteria 100% of the points pass the Gamma test for the uniform case, while 98.9% of the points do it for the lungs case.

  4. Commissioning of a novel microCT/RT system for small animal conformal radiotherapy

    NASA Astrophysics Data System (ADS)

    Rodriguez, Manuel; Zhou, Hu; Keall, Paul; Graves, Edward

    2009-06-01

    The purpose of this work was to commission a 120 kVp photon beam produced by a micro-computed tomography (microCT) scanner for use in irradiating mice to therapeutic doses. A variable-aperture collimator has been integrated with a microCT scanner to allow the delivery of beams with pseudocircular profiles of arbitrary width between 0.1 and 6.0 cm. The dose rate at the isocenter of the system was measured using ion chamber and gafchromic EBT film as 1.56-2.13 Gy min-1 at the water surface for field diameters between 0.2 and 6.0 cm. The dose rate decreases approximately 10% per every 5 mm depth in water for field diameters between 0.5 and 1.0 cm. The flatness, symmetry and penumbra of the beam are 3.6%, 1.0% and 0.5 mm, respectively. These parameters are sufficient to accurately conform the radiation dose delivered to target organs on mice. The irradiated field size is affected principally by the divergence of the beam. In general, the beam has appropriate dosimetric characteristics to accurately deliver the dose to organs inside the mice's bodies. Using multiple beams delivered from a variety of angular directions, targets as small as 2 mm may be irradiated while sparing surrounding tissue. This microCT/RT system is a feasible tool to irradiate mice using treatment planning and delivery methods analogous to those applied to humans.

  5. Planning and delivery comparison of six linac-based stereotactic radiosurgery techniques

    NASA Astrophysics Data System (ADS)

    Thakur, Varun Singh

    This work presents planning and delivery comparison of linac-based SRS treatment techniques currently available for single lesion cranial SRS. In total, two dedicated SRS systems (Novalis Tx, Cyberknife) and a HI-ART TomoTherapy system with six different delivery techniques are evaluated. Four delivery techniques are evaluated on a Novalis Tx system: circular cones, dynamic conformal arcs (DCA), static non-coplanar intensity modulated radiotherapy (NCP-IMRT), and volumetric modulated arc therapy (RapidArc) techniques are compared with intensity modulation based helical Tomotherapy on the HI-ART Tomotherapy system and with non-isocentric, multiple overlapping based robotic radiosurgery using the CyberKnife system. Thirteen patients are retrospectively selected for the study. The target volumes of each patient are transferred to a CT scan of a Lucy phantom (Standard Imaging Inc., Middleton, WI, USA) designed for end-to-end SRS QA. In order to evaluate the plans, several indices scoring the conformality, homogeneity and gradients in the plan are calculated and compared for each of the plans. Finally, to check the clinical deliverability of the plans and the delivery accuracy of different systems, a few targets are delivered on each system. A comparison between planned dose on treatment planning system and dose delivered on Gafchromic EBT film (ISP, Wayne, New Jersey, USA) is carried out by comparing dose beam profiles, isodose lines and by calculating gamma index.

  6. The Effect of Oblique Electron Beams to the Surface Dose Under the Bolus

    SciTech Connect

    Demir, Bayram Okutan, Murat; Cakir, Aydin; Goeksel, Evren; Bilge, Hatice

    2009-01-01

    The aim of this study is to determine the effect of bolus to the surface dose in oblique electron incidences. Irradiations with 4.5, 6, 7.5, 9, and 12-MeV electron beams were made for the incidence angles of 0 deg., 15 deg., 30 deg., 45 deg., 60 deg., 75 deg. and using 3 different bolus setups: (1) unbolused (no bolus), (2) 5-mm bolus, and (3) 10-mm bolus. A set of EBT gafchromic film pieces placed on the phantom surface was irradiated with a 400-cGy dose at D{sub max} for each setup. Whereas surface dose increased with increasing incidence degrees in the absence of a bolus, it was seen that there was a large surface dose decreasing in the presence of a bolus with increasing incidence angles. For 60 deg. incidence angle, the relative surface doses with unbolused setup were: 88.10%, 90.06%, 89.35%, 90.25%, and 97.10%; with 5-mm bolus: 66.45%, 81.20%, 99.78%, 124.43%, and 116.07%; and with 10-mm bolus: 22.65%, 45.20%, 55.20%, 65.82%, and 90.27% for 4.5, 6, 7.5, 9, and 12 MeV, respectively. The use of bolus in the treatment of highly oblique surfaces with low-energy electron beams significantly decreases the surface dose.

  7. Dosimetric impact of interplay effect in lung IMRT and VMAT treatment using in-house dynamic thorax phantom

    NASA Astrophysics Data System (ADS)

    Mukhlisin; Pawiro, S. A.

    2016-03-01

    Tumor motion due to patient's respiratory is a significant problem in radiotherapy treatment of lung cancer. The purpose of this project is to study the interplay effect through dosimetry verification between the calculated and delivered dose, as well as the dosimetric impact of leaf interplay with breathing-induced tumor motion in IMRT and VMAT treatment. In this study, a dynamic thorax phantom was designed and constructed for dosimetry measurement. The phantom had a linear sinusoidal tumor motion toward superior-inferior direction with variation of amplitudes and periods. TLD-100 LiF:Mg,Ti and Gafchromic EBT2 film were used to measure dose in the midpoint target and the spinal cord. The IMRT and VMAT treatment had prescription dose of 200 cGy per fraction. The dosimetric impact due to interplay effect during IMRT and VMAT treatment were resulted in the range of 0.5% to -6.6% and 0.9% to -5.3% of target dose reduction, respectively. Meanwhile, mean dose deviation of spinal cord in IMRT and VMAT treatment were around 1.0% to -6.9% and 0.9% to -6.3%, respectively. The results showed that if respiratory management technique were not implemented, the presence of lung tumor motion during dose delivery in IMRT and VMAT treatment causes dose discrepancies inside tumor volume.

  8. γ-H2AX and phosphorylated ATM focus formation in cancer cells after laser plasma X irradiation.

    PubMed

    Sato, Katsutoshi; Nishikino, Masaharu; Okano, Yasuaki; Ohshima, Shinsuke; Hasegawa, Noboru; Ishino, Masahiko; Kawachi, Tetsuya; Numasaki, Hodaka; Teshima, Teruki; Nishimura, Hiroaki

    2010-10-01

    The usefulness of laser plasma X-ray pulses for medical and radiation biological studies was investigated, and the effects of laser plasma X rays were compared with those of conventional sources such as a linear accelerator. A cell irradiation system was developed that used copper-Kα (8 keV) lines from an ultrashort high-intensity laser to produce plasma. The absorbed dose of the 8 keV laser plasma X-ray pulse was estimated accurately with Gafchromic® EBT film. When the cells were irradiated with approximately 2 Gy of laser plasma X rays, the circular regions on γ-H2AX-positive cells could be clearly identified. Moreover, the numbers of γ-H2AX and phosphorylated ataxia telangiectasia mutated (ATM) foci induced by 8 keV laser plasma X rays were comparable to those induced by 4 MV X rays. These results indicate that the laser plasma X ray source may be useful for radiation biology studies. PMID:20718602

  9. Use of 3D printers to create a patient-specific 3D bolus for external beam therapy.

    PubMed

    Burleson, Sarah; Baker, Jamie; Hsia, An Ting; Xu, Zhigang

    2015-01-01

    The purpose of this paper is to demonstrate that an inexpensive 3D printer can be used to manufacture patient-specific bolus for external beam therapy, and to show we can accurately model this printed bolus in our treatment planning system for accurate treatment delivery. Percent depth-dose measurements and tissue maximum ratios were used to determine the characteristics of the printing materials, acrylonitrile butadiene styrene and polylactic acid, as bolus material with physical density of 1.04 and 1.2 g/cm3, and electron density of 3.38 × 10²³ electrons/cm3 and 3.80 × 10²³ electrons/ cm3, respectively. Dose plane comparisons using Gafchromic EBT2 film and the RANDO phantom were used to verify accurate treatment planning. We accurately modeled a printing material in Eclipse treatment planning system, assigning it a Hounsfield unit of 260. We were also able to verify accurate treatment planning using gamma analysis for dose plane comparisons. With gamma criteria of 5% dose difference and 2 mm DTA, we were able to have 86.5% points passing, and with gamma criteria of 5% dose difference and 3 mm DTA, we were able to have 95% points passing. We were able to create a patient-specific bolus using an inexpensive 3D printer and model it in our treatment planning system for accurate treatment delivery. PMID:26103485

  10. Comparison of selected dose calculation algorithms in radiotherapy treatment planning for tissues with inhomogeneities

    NASA Astrophysics Data System (ADS)

    Woon, Y. L.; Heng, S. P.; Wong, J. H. D.; Ung, N. M.

    2016-03-01

    Inhomogeneity correction is recommended for accurate dose calculation in radiotherapy treatment planning since human body are highly inhomogeneous with the presence of bones and air cavities. However, each dose calculation algorithm has its own limitations. This study is to assess the accuracy of five algorithms that are currently implemented for treatment planning, including pencil beam convolution (PBC), superposition (SP), anisotropic analytical algorithm (AAA), Monte Carlo (MC) and Acuros XB (AXB). The calculated dose was compared with the measured dose using radiochromic film (Gafchromic EBT2) in inhomogeneous phantoms. In addition, the dosimetric impact of different algorithms on intensity modulated radiotherapy (IMRT) was studied for head and neck region. MC had the best agreement with the measured percentage depth dose (PDD) within the inhomogeneous region. This was followed by AXB, AAA, SP and PBC. For IMRT planning, MC algorithm is recommended for treatment planning in preference to PBC and SP. The MC and AXB algorithms were found to have better accuracy in terms of inhomogeneity correction and should be used for tumour volume within the proximity of inhomogeneous structures.

  11. Experimental and Monte Carlo measurements of dose perturbation around a non-radioactive brachytherapy seed in external beam radiotherapy

    NASA Astrophysics Data System (ADS)

    Steinman, James P.

    I-125 seeds used in permanent prostate brachytherapy are composed of high-Z metals and may number from 40 to over 100 in a typical implant. If any supplemental external beam treatment is administered afterward (as for salvaging failed brachytherapy treatment), it is possible that the seeds may cause substantial dose perturbation which will depend on numerous factors (photon energy, depth, field size, number of seeds, etc.) and this effect needs to be thoroughly investigated. Film measurements were primarily done using Kodak XV2 layered above and below a non-radioactive I-125 seed placed in a groove on a Lucite plate with 5 cm buildup and 10 cm backscatter added at 95 cm SSD. The phantom was irradiated with and without seed with 6 MV photons for a 1 x 1 cm2 field size. Monte Carlo simulations were carried out using DOSXYZnrc using the same parameters and compared with Gafchromic EBT2 film. Other comparisons looked at changing energy, depth, and field size in both with and without seeds configuration. This study was further extended to include metals of various Z of the seed's dimensions and also looked into effect of 3 seeds spaced 0.5 cm vertically. Another measurement was done using two opposing fields using single as well as 3 seed configuration to see whether the dose enhancement and attenuation cancel out in multi-field treatments which is the norm clinically in a prostate treatment. For a single I-125 seed, on XV film a localized dose enhancement of 6.3% upstream and -10.9% downstream was noticed. With three seeds, this effect did not change. With two opposing fields, a cold spot around the seed of ~3% was noticed from film measurements. Increasing energy and field size decreased the effect while increase in Z of material greatly increased the effect. Increasing depth appeared to have no effect. DOSXYZnrc and EBT2 film verified maximum dose enhancement of +15% upstream and -20% downstream of the I-125 seed surface. In general, the range of the effect was

  12. Films and Film Sources for Materials Science and Engineering Courses

    ERIC Educational Resources Information Center

    Evans, Edward B.

    1972-01-01

    A selected list of films that are suitable for secondary schools through universities. They cover all phases of materials science and engineering. The films may be obtained, usually free of charge, for listed sources. (DF)

  13. SU-E-T-384: Experimental Verification of a Monte Carlo Linear Accelerator Model Using a Radiochromic Film Stack Dosimeter

    SciTech Connect

    McCaw, T; Culberson, W; DeWerd, L

    2014-06-01

    Purpose: To experimentally verify a Monte Carlo (MC) linear accelerator model for the simulation of intensity-modulated radiation therapy (IMRT) treatments of moving targets. Methods: A Varian Clinac™ 21EX linear accelerator was modeled using the EGSnrc user code BEAMnrc. The mean energy, radial-intensity distribution, and divergence of the electron beam incident on the bremsstrahlung target were adjusted to achieve agreement between simulated and measured percentage-depth-dose and transverse field profiles for a 6 MV beam. A seven-field step-and-shoot IMRT lung procedure was prepared using Varian Eclipse™ treatment planning software. The plan was delivered using a Clinac™ 21EX linear accelerator and measured with a Gafchromic™ EBT2 film stack dosimeter (FSD) in two separate static geometries: within a cylindrical water-equivalent-plastic phantom and within an anthropomorphic chest phantom. Two measurements were completed in each setup. The dose distribution for each geometry was simulated using the EGSnrc user code DOSXYZnrc. MC geometries of the treatment couch, cylindrical phantom, and chest phantom were developed by thresholding CT data sets using MATLAB™. The FSD was modeled as water. The measured and simulated dose distributions were normalized to the median dose within the FSD. Results: Using an electron beam with a mean energy of 6.05 MeV, a Gaussian radial-intensity distribution with a full width at half maximum of 1.5 mm, and a divergence of 0°, the measured and simulated dose profiles agree within 1.75% and 1 mm. Measured and simulated dose distributions within both the cylindrical and chest phantoms agree within 3% over 94% of the FSD volume. The overall uncertainty in the FSD measurements is 3.1% (k=1). Conclusion: MC simulations agree with FSD measurements within measurement uncertainty, thereby verifying the accuracy of the linear accelerator model for the simulation of IMRT treatments of static geometries. The experimental verification

  14. On the suitability of ultrathin detectors for absorbed dose assessment in the presence of high-density heterogeneities

    SciTech Connect

    Bueno, M. Duch, M. A.; Carrasco, P.; Jornet, N.; Muñoz-Montplet, C.

    2014-08-15

    Purpose: The aim of this study was to evaluate the suitability of several detectors for the determination of absorbed dose in bone. Methods: Three types of ultrathin LiF-based thermoluminescent dosimeters (TLDs)—two LiF:Mg,Cu,P-based (MCP-Ns and TLD-2000F) and a{sup 7}Li-enriched LiF:Mg,Ti-based (MTS-7s)—as well as EBT2 Gafchromic films were used to measure percentage depth-dose distributions (PDDs) in a water-equivalent phantom with a bone-equivalent heterogeneity for 6 and 18 MV and a set of field sizes ranging from 5×5 cm{sup 2} to 20×20 cm{sup 2}. MCP-Ns, TLD-2000F, MTS-7s, and EBT2 have active layers of 50, 20, 50, and 30 μm, respectively. Monte Carlo (MC) dose calculations (PENELOPE code) were used as the reference and helped to understand the experimental results and to evaluate the potential perturbation of the fluence in bone caused by the presence of the detectors. The energy dependence and linearity of the TLDs’ response was evaluated. Results: TLDs exhibited flat energy responses (within 2.5%) and linearity with dose (within 1.1%) within the range of interest for the selected beams. The results revealed that all considered detectors perturb the electron fluence with respect to the energy inside the bone-equivalent material. MCP-Ns and MTS-7s underestimated the absorbed dose in bone by 4%–5%. EBT2 exhibited comparable accuracy to MTS-7s and MCP-Ns. TLD-2000F was able to determine the dose within 2% accuracy. No dependence on the beam energy or field size was observed. The MC calculations showed that a50 μm thick detector can provide reliable dose estimations in bone regardless of whether it is made of LiF, water or EBT’s active layer material. Conclusions: TLD-2000F was found to be suitable for providing reliable absorbed dose measurements in the presence of bone for high-energy x-ray beams.

  15. Epitaxial thin films

    DOEpatents

    Hunt, Andrew Tye; Deshpande, Girish; Lin, Wen-Yi; Jan, Tzyy-Jiuan

    2006-04-25

    Epitatial thin films for use as buffer layers for high temperature superconductors, electrolytes in solid oxide fuel cells (SOFC), gas separation membranes or dielectric material in electronic devices, are disclosed. By using CCVD, CACVD or any other suitable deposition process, epitaxial films having pore-free, ideal grain boundaries, and dense structure can be formed. Several different types of materials are disclosed for use as buffer layers in high temperature superconductors. In addition, the use of epitaxial thin films for electrolytes and electrode formation in SOFCs results in densification for pore-free and ideal gain boundary/interface microstructure. Gas separation membranes for the production of oxygen and hydrogen are also disclosed. These semipermeable membranes are formed by high-quality, dense, gas-tight, pinhole free sub-micro scale layers of mixed-conducting oxides on porous ceramic substrates. Epitaxial thin films as dielectric material in capacitors are also taught herein. Capacitors are utilized according to their capacitance values which are dependent on their physical structure and dielectric permittivity. The epitaxial thin films of the current invention form low-loss dielectric layers with extremely high permittivity. This high permittivity allows for the formation of capacitors that can have their capacitance adjusted by applying a DC bias between their electrodes.

  16. Thin film scintillators

    NASA Astrophysics Data System (ADS)

    McDonald, Warren; McKinney, George; Tzolov, Marian

    2015-03-01

    Scintillating materials convert energy flux (particles or electromagnetic waves) into light with spectral characteristic matching a subsequent light detector. Commercial scintillators such as yttrium aluminum garnet (YAG) and yttrium aluminum perovskite (YAP) are commonly used. These are inefficient at lower energies due to the conductive coating present on their top surface, which is needed to avoid charging. We hypothesize that nano-structured thin film scintillators will outperform the commercial scintillators at low electron energies. We have developed alternative thin film scintillators, zinc tungstate and zinc oxide, which show promise for higher sensitivity to lower energy electrons since they are inherently conductive. Zinc tungstate films exhibit photoluminescence quantum efficiency of 74%. Cathodoluminescence spectroscopy was applied in transmission and reflection geometries. The comparison between the thin films and the YAG and YAP commercial scintillators shows much higher light output from the zinc tungstate and zinc oxide at electron energies less than 5 keV. Our films were integrated in a backscattered electron detector. This detector delivers better images than an identical detector with commercial YAG scintillator at low electron energies. Dr. Nicholas Barbi from PulseTor LLC, Dr. Anura Goonewardene, NSF Grants: #0806660, #1058829, #0923047.

  17. Sensitometric properties of radiographic films and film classification

    NASA Astrophysics Data System (ADS)

    Siidorow, A.

    An attempt was made to elucidate those film characteristics which affect the radiographic image quality and which could be used as the basis of film classification. A literary survey showed that it is apparently impossible to measure the most important film characteristics, i.e., graininess and granularity, in a reliable and simple way. An existing classification method based on a visual comparison of film graininess is recommended.

  18. From Written Film History to Visual Film History.

    ERIC Educational Resources Information Center

    Petric, Vladimir

    The poor quality of most university courses in film history is due to several factors, among them the fact that there is insufficient analytical documentation and direct cinematic illustration in existent written film histories. These histories examine films on a thematic level, offering noncinematic interpretation such as literary meaning, social…

  19. Depositing Adherent Ag Films On Ti Films On Alumina

    NASA Technical Reports Server (NTRS)

    Honecy, Frank S.

    1995-01-01

    Report discusses cleaning of ceramic (principally, alumina) substrates in preparation for sputter deposition of titanium intermediate films on substrates followed by sputter deposition of outer silver films. Principal intended application, substrates sliding parts in advanced high-temperature heat engines, and outer silver films serve as solid lubricants: lubricating properties described in "Solid Lubricant for Alumina" (LEW-15495).

  20. The New Film Technologies: Computerized Video-Assisted Film Production.

    ERIC Educational Resources Information Center

    Mott, Donald R.

    Over the past few years, video technology has been used to assist film directors after they have shot a scene, to control costs, and to create special effects, especially computer assisted graphics. At present, a computer based editing system called "Film 5" combines computer technology and video tape with film to save as much as 50% of the cost…

  1. A multilayer sonic film

    NASA Astrophysics Data System (ADS)

    Munteanu, L.; Chiroiu, V.; Sireteanu, T.; Dumitriu, D.

    2015-10-01

    A non-periodic multilayer film was analyzed to show that, despite its non-periodicity, the film exhibits full band-gaps and localized modes at its interfaces, as well as in the sonic composites. The film consists of alternating layers of two different materials that follow a triadic Cantor sequence. The Cantor structure shows extremely low thresholds for subharmonic generation of ultrasonic waves, compared with homogeneous and periodic structures. The coupling between the extended-mode (phonon) and the localized-mode (fracton) vibration regimes explains the generation of full band-gaps, for which there are no propagating Lamb waves. The large enhancement of the nonlinear interaction results from a more favorable frequency and spatial matching of coupled modes. A full band-gap that excludes Love waves is also analyzed.

  2. Thin film temperature sensor

    NASA Technical Reports Server (NTRS)

    Grant, H. P.; Przybyszewski, J. S.

    1980-01-01

    Thin film surface temperature sensors were developed. The sensors were made of platinum-platinum/10 percent rhodium thermocouples with associated thin film-to-lead wire connections and sputtered on aluminum oxide coated simulated turbine blades for testing. Tests included exposure to vibration, low velocity hydrocarbon hot gas flow to 1250 K, and furnace calibrations. Thermal electromotive force was typically two percent below standard type S thermocouples. Mean time to failure was 42 hours at a hot gas flow temperature of 1250 K and an average of 15 cycles to room temperature. Failures were mainly due to separation of the platinum thin film from the aluminum oxide surface. Several techniques to improve the adhesion of the platinum are discussed.

  3. Photographic film image enhancement

    NASA Technical Reports Server (NTRS)

    Horner, J. L.

    1975-01-01

    A series of experiments were undertaken to assess the feasibility of defogging color film by the techniques of optical spatial filtering. A coherent optical processor was built using red, blue, and green laser light input and specially designed Fourier transformation lenses. An array of spatial filters was fabricated on black and white emulsion slides using the coherent optical processor. The technique was first applied to laboratory white light fogged film, and the results were successful. However, when the same technique was applied to some original Apollo X radiation fogged color negatives, the results showed no similar restoration. Examples of each experiment are presented and possible reasons for the lack of restoration in the Apollo films are discussed.

  4. Thin film photovoltaics

    SciTech Connect

    Zweibel, K; Ullal, H S

    1989-05-01

    Thin films are considered a potentially attractive technological approach to making cost-effective electricity by photovoltaics. Over the last twenty years, many have been investigated and some (cadmium telluride, copper indium diselenide, amorphous silicon) have become leading candidates for future large-scale commercialization. This paper surveys the past development of these key thin films and gives their status and future prospects. In all cases, significant progress toward cost-effective PV electricity has been made. If this progress continues, it appears that thin film PV could provide electricity that is competitive for summer daytime peaking power requirements by the middle of the 1990s; and electricity in a range that is competitive with fossil fuel costs (i.e., 6 cents/kilowatt-hour) should be available from PV around the turn of the century. 22 refs., 9 figs.

  5. Thin films and uses

    DOEpatents

    Baskaran, Suresh; Graff, Gordon L.; Song, Lin

    1998-01-01

    The invention provides a method for synthesizing a titanium oxide-containing film comprising the following steps: (a) preparing an aqueous solution of a titanium chelate with a titanium molarity in the range of 0.01M to 0.6M. (b) immersing a substrate in the prepared solution, (c) decomposing the titanium chelate to deposit a film on the substrate. The titanium chelate maybe decomposed acid, base, temperature or other means. A preferred method provides for the deposit of adherent titanium oxide films from C2 to C5 hydroxy carboxylic acids. In another aspect the invention is a novel article of manufacture having a titanium coating which protects the substrate against ultraviolet damage. In another aspect the invention provides novel semipermeable gas separation membranes, and a method for producing them.

  6. Foundation for Film and Science

    ERIC Educational Resources Information Center

    van der Veen, G.

    1976-01-01

    Provides a comprehensive discussion on the Stichting Film en Wetenschap, SFW (Foundation for Film and Science), in Utrecht. Various aspects of the use of audio-visual aids in university teaching are looked at in detail. (Editor/RK)

  7. Thin films for material engineering

    NASA Astrophysics Data System (ADS)

    Wasa, Kiyotaka

    2016-07-01

    Thin films are defined as two-dimensional materials formed by condensing one by one atomic/molecular/ionic species of matter in contrast to bulk three-dimensional sintered ceramics. They are grown through atomic collisional chemical reaction on a substrate surface. Thin film growth processes are fascinating for developing innovative exotic materials. On the basis of my long research on sputtering deposition, this paper firstly describes the kinetic energy effect of sputtered adatoms on thin film growth and discusses on a possibility of room-temperature growth of cubic diamond crystallites and the perovskite thin films of binary compound PbTiO3. Secondly, high-performance sputtered ferroelectric thin films with extraordinary excellent crystallinity compatible with MBE deposited thin films are described in relation to a possible application for thin-film MEMS. Finally, the present thin-film technologies are discussed in terms of a future material science and engineering.

  8. The Nuclear Debate in Film

    ERIC Educational Resources Information Center

    Dowling, John

    1977-01-01

    Provides a nuclear film bibliography grouped into the areas of: building and using the bomb; living with the bomb; and living with nuclear power. These films are for mature high school students and older. (MLH)

  9. Thin film solar cell workshop

    NASA Technical Reports Server (NTRS)

    Armstrong, Joe; Jeffrey, Frank

    1993-01-01

    A summation of responses to questions posed to the thin-film solar cell workshop and the ensuing discussion is provided. Participants in the workshop included photovoltaic manufacturers (both thin film and crystalline), cell performance investigators, and consumers.

  10. Thin film ceramic thermocouples

    NASA Technical Reports Server (NTRS)

    Gregory, Otto (Inventor); Fralick, Gustave (Inventor); Wrbanek, John (Inventor); You, Tao (Inventor)

    2011-01-01

    A thin film ceramic thermocouple (10) having two ceramic thermocouple (12, 14) that are in contact with each other in at least on point to form a junction, and wherein each element was prepared in a different oxygen/nitrogen/argon plasma. Since each element is prepared under different plasma conditions, they have different electrical conductivity and different charge carrier concentration. The thin film thermocouple (10) can be transparent. A versatile ceramic sensor system having an RTD heat flux sensor can be combined with a thermocouple and a strain sensor to yield a multifunctional ceramic sensor array. The transparent ceramic temperature sensor that could ultimately be used for calibration of optical sensors.

  11. Thin film photovoltaic device

    DOEpatents

    Catalano, Anthony W.; Bhushan, Manjul

    1982-01-01

    A thin film photovoltaic solar cell which utilizes a zinc phosphide semiconductor is of the homojunction type comprising an n-type conductivity region forming an electrical junction with a p-type region, both regions consisting essentially of the same semiconductor material. The n-type region is formed by treating zinc phosphide with an extrinsic dopant such as magnesium. The semiconductor is formed on a multilayer substrate which acts as an opaque contact. Various transparent contacts may be used, including a thin metal film of the same chemical composition as the n-type dopant or conductive oxides or metal grids.

  12. Sepis education on film.

    PubMed

    2016-09-01

    Early identification and management is vital in treating sepsis, and a new film, which aims to help clinicians spot and respond to warning signs of sepsis in children, has been launched by NHS Health Education England. The film, which features parents who lost their young daughter to undiagnosed sepsis, highlights the important signs clinicians should look for, and asks them to think 'could this be sepsis' when assessing children. It also highlights the wide range of learning resources available, including the new National Institute of Health and Care Excellence guidelines ( nice.org.uk/guidance/NG51 ). PMID:27581900

  13. Thin film photovoltaic device

    DOEpatents

    Catalano, A.W.; Bhushan, M.

    1982-08-03

    A thin film photovoltaic solar cell which utilizes a zinc phosphide semiconductor is of the homojunction type comprising an n-type conductivity region forming an electrical junction with a p-type region, both regions consisting essentially of the same semiconductor material. The n-type region is formed by treating zinc phosphide with an extrinsic dopant such as magnesium. The semiconductor is formed on a multilayer substrate which acts as an opaque contact. Various transparent contacts may be used, including a thin metal film of the same chemical composition as the n-type dopant or conductive oxides or metal grids. 5 figs.

  14. Current Film Periodicals in English. Revised Edition.

    ERIC Educational Resources Information Center

    Reilly, Adam, Comp.

    This bibliography of about 200 periodicals dealing with film covers several types of magazine: scholarly journals on film aesthetics, like "The Film Journal"; news notes for movie fans, like "Film Nut News"; magazines which cover films as well as the other arts, like "Cue" and "After Dark"; film education periodicals, like "Media and Methods";…

  15. Filming in decontamination by mopping

    SciTech Connect

    Rankin, W.N.; Toole, P.A.

    1993-09-28

    Technical assistance was provided High Level Waste Engineering in the investigation and prevention of filming during decontamination by mopping. After mopping operations in a Tank Farm application, a film of the cleaning agent sometimes remained on the surface being cleaned which interfered with monitoring to detect the presence of radioactive material. Scoping tests were conducted to investigate filming characteristics of two cleaning materials. In addition, rinsing test were conducted to demonstrate how filming can be prevented.

  16. Dual clearance squeeze film damper

    NASA Technical Reports Server (NTRS)

    Fleming, D. P. (Inventor)

    1985-01-01

    A dual clearance hydrodynamic liquid squeeze film damper for a gas turbine engine is described. Under normal operating conditions, the device functions as a conventional squeeze film damper, using only one of its oil films. When an unbalance reaches abusive levels, as may occur with a blade loss or foreign object damage, a second, larger clearance film becomes active, controlling vibration amplitudes in a near optimum manner until the engine can be safely shut down and repaired.

  17. NMR characterization of thin films

    DOEpatents

    Gerald II, Rex E.; Klingler, Robert J.; Rathke, Jerome W.; Diaz, Rocio; Vukovic, Lela

    2010-06-15

    A method, apparatus, and system for characterizing thin film materials. The method, apparatus, and system includes a container for receiving a starting material, applying a gravitational force, a magnetic force, and an electric force or combinations thereof to at least the starting material, forming a thin film material, sensing an NMR signal from the thin film material and analyzing the NMR signal to characterize the thin film of material.

  18. NMR characterization of thin films

    DOEpatents

    Gerald, II, Rex E.; Klingler, Robert J.; Rathke, Jerome W.; Diaz, Rocio; Vukovic, Lela

    2008-11-25

    A method, apparatus, and system for characterizing thin film materials. The method, apparatus, and system includes a container for receiving a starting material, applying a gravitational force, a magnetic force, and an electric force or combinations thereof to at least the starting material, forming a thin film material, sensing an NMR signal from the thin film material and analyzing the NMR signal to characterize the thin film of material.

  19. Instructional Films: Asset or Liability?

    ERIC Educational Resources Information Center

    Braselman, Herbert P.

    1978-01-01

    The unique capabilities of film and research findings in educational psychology, learning psychology, and cost effectiveness indicate that film is an asset to the educational process. Sufficient resources and continued training should be provided to enable teachers to use the most effective film at the optimal time. (CMV)

  20. Film Images of the Negro.

    ERIC Educational Resources Information Center

    Manchel, Frank

    1967-01-01

    Educators can help students recognize the value of the motion picture as a social influence by exposing them to film stereotyping and the effect of this distortion on society. A historical study of the film image of the Negro will show him emerging from a humorous, fearful, "perverted" character in early films to "an unfortunate member of society"…

  1. Film Study in Higher Education.

    ERIC Educational Resources Information Center

    Stewart, David C., Ed.

    This resource book on the place of film study today is designed to assist in the planning of college courses in the history, criticism, and appreciation of motion pictures. Representative course descriptions and appraisals are given by (1) Jack C. Ellis, who describes a two-part course in film aesthetics and types of films, (2) Edward Fischer, who…

  2. Radical Pedagogy, Prison, and Film

    ERIC Educational Resources Information Center

    O'Neill, Dierdre

    2015-01-01

    This article explores the work of The Inside Film project. Inside Film works with a specific group of people (prisoners and ex-prisoners) in a particular set of circumstances (in prison or on parole) exploring how film making can be used within prison education or with people who have been to prison as a means of fostering a critical engagement…

  3. Nonfiction Film Theory and Criticism.

    ERIC Educational Resources Information Center

    Barsam, Richard Meran, Ed.

    This anthology offers significant writings on the subject of nonfiction film, with emphasis primarily on British and American films and film makers (although the work of Alberto Cavalcanti, Leni Riefenstahl, and Joris Ivens is also represented). The first section of the book offers essays that examine and define the essential nature and idea of…

  4. Archaeology on Film. Second Edition.

    ERIC Educational Resources Information Center

    Downs, Mary, Ed.; And Others

    This document provides a comprehensive guide to archaeological films and video tapes of archaeological interest. Individual films and film series are listed alphabetically by title. Each entry includes the following information: title, series, date, length, color/black & white, format, purchase and rental prices, distributor/rental source,…

  5. Longevity Of Dry Film Lubricants

    NASA Technical Reports Server (NTRS)

    Kannel, J. W.; Stockwell, R. D.

    1993-01-01

    Report describes evaluation of dry film lubricants candidate for use in rotary joints of proposed Space Station. Study included experiments and theoretical analyses focused on longevity of sputtered molybdenum disulfide films and ion-plated lead films under conditions partially simulating rolling contact.

  6. Teaching Argumentative Writing through Film.

    ERIC Educational Resources Information Center

    Fluitt-Dupuy, Jan

    2001-01-01

    Discusses how watching and discussing feature films and writing reviews of these films in the English-as-a-Second/Foreign-Language classroom can be instrumental in teaching the principles of good argumentative writing within the confines of the simple movie review. Six steps for teaching a film review unit are provided. (Author/VWL)

  7. Designing and Dosimetry of a Shield for Photon Fields of Radiation Therapy in Oral Cavity Cancer

    PubMed Central

    Jabbari, Keyvan; Senobari, Somayeh; Roayaei, Mahnaz; Rostampour, Masoumeh

    2015-01-01

    The cancer of oral cavity is related to lesions of mucous membrane of tongue and gum that can be treated with radiation therapy. A lateral photon field can be used to treat this kind of tumor, which has a side-effect on normal tissue in the opposite side of the oral cavity. In this study the dosimetric effect of the various shields in oral cavity is evaluated. In this study, a special phantom similar to the structure of oral cavity with capability of film dosimetry was designed and constructed. The various shield slabs were made of five materials: Lead, Plexiglas, Acrylic resin, Silicon and Plaster. For irradiation, Cobalt 60 (60Co) and 6 MV photon beams were used. The film dosimetry before and after the shield was performed using GAFCHROMIC EBT2 films. The film before the shield measures the magnitude of backscattering radiation from the shield. The prescribed dose was 150 cGy. Results showed that 3 cm of the lead in both energies had the maximum absorption of radiation. The absorbed dose to opposite side of shield for 6 MV photon beams and 60Co were 21 and 32 cGy, respectively. The minimum attenuation on radiation was observed in silicon shield for which the dose of opposite side were 116 and 147 cGy for 6 MV and 60Co respectively. The maximum backscattered dose was measured 177 cGy and 219 cGy using 3 cm thickness of lead, which was quite considerable. The minimum backscattering where for acrylic resin 101 and 118 cGy for 6 MV and cobalt. In this study, it was concluded that the amount of backscattering for 3 cm Lead shield is quite considerable and increases the dose significantly. A composite layer of shield with 1–2 cm lead and 1 cm acrylic resin can have the protective effect and low backscattering radiation at the same time. PMID:26120570

  8. Designing and Dosimetry of a Shield for Photon Fields of Radiation Therapy in Oral Cavity Cancer.

    PubMed

    Jabbari, Keyvan; Senobari, Somayeh; Roayaei, Mahnaz; Rostampour, Masoumeh

    2015-01-01

    The cancer of oral cavity is related to lesions of mucous membrane of tongue and gum that can be treated with radiation therapy. A lateral photon field can be used to treat this kind of tumor, which has a side-effect on normal tissue in the opposite side of the oral cavity. In this study the dosimetric effect of the various shields in oral cavity is evaluated. In this study, a special phantom similar to the structure of oral cavity with capability of film dosimetry was designed and constructed. The various shield slabs were made of five materials: Lead, Plexiglas, Acrylic resin, Silicon and Plaster. For irradiation, Cobalt 60 (60Co) and 6 MV photon beams were used. The film dosimetry before and after the shield was performed using GAFCHROMIC EBT2 films. The film before the shield measures the magnitude of backscattering radiation from the shield. The prescribed dose was 150 cGy. Results showed that 3 cm of the lead in both energies had the maximum absorption of radiation. The absorbed dose to opposite side of shield for 6 MV photon beams and 60Co were 21 and 32 cGy, respectively. The minimum attenuation on radiation was observed in silicon shield for which the dose of opposite side were 116 and 147 cGy for 6 MV and 60Co respectively. The maximum backscattered dose was measured 177 cGy and 219 cGy using 3 cm thickness of lead, which was quite considerable. The minimum backscattering where for acrylic resin 101 and 118 cGy for 6 MV and cobalt. In this study, it was concluded that the amount of backscattering for 3 cm Lead shield is quite considerable and increases the dose significantly. A composite layer of shield with 1-2 cm lead and 1 cm acrylic resin can have the protective effect and low backscattering radiation at the same time. PMID:26120570

  9. SU-E-T-104: Development of 3 Dimensional Dosimetry System for Gamma Knife

    SciTech Connect

    Yoon, K; Kwak, J; Cho, B; Lee, D; Ahn, S

    2014-06-01

    Purpose: The aim of this study was to develop a new 3 dimensional dosimetry system to verify the dosimetric accuracy of Leksell Gamma Knife-Perfexion™ (LGK) (Elekta, Norcross, GA). Methods: We designed and manufactured a lightweight dosimetry instrument to be equipped with the head frame to LGK. It consists of a head phantom, a scintillator, a CCD camera and a step motor. The 10×10 cm2 sheet of Gd2O3;Tb phosphor or Gafchromic EBT3 film was located at the center of the 16 cm diameter hemispherical PMMA, the head phantom. The additional backscatter compensating material of 1 cm thick PMMA plate was placed downstream of the phosphor sheet. The backscatter plate was transparent for scintillation lights to reach the CCD camera with 1200×1200 pixels by 5.2 um pitch. With This equipment, 300 images with 0.2 mm of slice gap were acquired under three collimator setups (4mm, 8mm and 16mm), respectively. The 2D projected doses from 3D distributions were compared with the exposured film dose. Results: As all doses normalized by the maximum dose value in 16 mm setup, the relative differences between the equipment dose and film dose were 0.2% for 4mm collimator and 0.5% for 8mm. The acquisition of 300 images by the equipment took less than 3 minutes. Conclusion: The new equipment was verified to be a good substitute to radiochromic film, with which required more time and resources. Especially, the new methods was considered to provide much convenient and faster solution in the 3D dose acquisition for LGK.

  10. SU-E-T-446: Evaluation of the Dosimetric Properties of a Diode Detector to Proton Radiosurgery

    SciTech Connect

    Teran, A; McAuley, G; Slater, J M; Slater, J D; Wroe, A

    2014-06-01

    Purpose: To test the PTW PR60020 proton dosimetry diode in radiation fields relevant to proton radiosurgery applications and evaluate its suitability as a high resolution, real time dosimetry device. Methods: Data was collected using our standard nominal radiosurgery energies of 126 MeV and 155 MeV through a single stage scattering system, corresponding to a range of 9.7 and 15 cm in water respectively. Various beam modulations were tested as part of this study. Depth dose and beam profile measurements were completed with the PTW PR60020 dosimetry diode with comparative measurements using a PTW Markus ionization chamber and EBT2 Gafchromic film. Monte Carlo simulations were also completed for comparison. Results: The single 1 mm{sup 2} by 20 μm thick sensitive volume allowed for high spatial resolution measurements while maintaining sufficient sensitive volume to ensure that measurements could be completed without excessive beam delivery. Depth dose profiles exhibited negligible LET dependence which typically impacts film and other solid state dosimetry devices, while beam ranges measured with the PTW diode were within 1 mm of ion chamber data. In an edge on arrangement beam profiles were also measured within 0.5 mm full-width at half-maximum at various depths as compared to film and simulation data. Conclusion: The PTW PR60020 proved to be a very useful radiation metrology apparatus for proton radiosurgery applications. Its waterproof and rugged construction allowed for easy deployment in phantoms or water tanks that are commonly used in proton radiosurgery QA. Dosimetrically, the diode exhibited negligible LET dependence as a function of depth, while in edge on arrangement to the incident proton beam it facilitated the measurement of beam profiles with a spatial resolution comparable to both Monte Carlo and film measurements. This project was sponsored in part by funding from the Department of Defense (DOD# W81XWH-BAA-10-1)

  11. Methods for producing complex films, and films produced thereby

    DOEpatents

    Duty, Chad E.; Bennett, Charlee J. C.; Moon, Ji -Won; Phelps, Tommy J.; Blue, Craig A.; Dai, Quanqin; Hu, Michael Z.; Ivanov, Ilia N.; Jellison, Jr., Gerald E.; Love, Lonnie J.; Ott, Ronald D.; Parish, Chad M.; Walker, Steven

    2015-11-24

    A method for producing a film, the method comprising melting a layer of precursor particles on a substrate until at least a portion of the melted particles are planarized and merged to produce the film. The invention is also directed to a method for producing a photovoltaic film, the method comprising depositing particles having a photovoltaic or other property onto a substrate, and affixing the particles to the substrate, wherein the particles may or may not be subsequently melted. Also described herein are films produced by these methods, methods for producing a patterned film on a substrate, and methods for producing a multilayer structure.

  12. Museums With Film Programs.

    ERIC Educational Resources Information Center

    Melton, Hollis, Comp.

    Selectively listed by state are the names and addresses of museums in the U.S., Canada, and Puerto Rico which feature film programs as part of their regular activities. The list, acknowledged to be less than complete, includes some natural history and science museums, some anthropology museums, and state and county historical societies. The…

  13. Mobile Library Filming Device.

    ERIC Educational Resources Information Center

    Martin, Claud E.

    This report contains details of the study and performance test of the Mobile Filming Library Device which consists of a camera and self contained power source. Because of the cost savings and service improvement characteristics, this technique involving the use of a microfilm intermediate in the preparation of copies of material filed in full size…

  14. Film, Radio, and Television.

    ERIC Educational Resources Information Center

    Hardesty, Carolyn, Ed.

    1990-01-01

    This journal issue covers the history of film, radio, and television in Iowa. The first article, "When Pictures and Sound Came to Iowa," summarizes the origin of movies and radio and their early beginnings in Iowa. Using old photographs and measurement charts, the viewing, reading, and listening habits of young people in 1950 and 1958 are…

  15. Intercultural Training with Films

    ERIC Educational Resources Information Center

    Roell, Christine

    2010-01-01

    Films are a great medium to use not only to practice English, but also to facilitate intercultural learning. Today English is a global language spoken by people from many countries and cultural backgrounds. Since culture greatly impacts communication, it is helpful for teachers to introduce lessons and activities that reveal how different…

  16. Film: The Creative Eye.

    ERIC Educational Resources Information Center

    Sohn, David A.

    Short films are often experimental in nature. They can place aspects of the environment which are usually unnoticed in such a way as to sharpen our observations of the world, and "create a new awareness, a fuller sense of life and being." Based on the premise that visual literacy is becoming increasingly important, this book describes several…

  17. Paradoxes in Film Ratings

    ERIC Educational Resources Information Center

    Moore, Thomas L.

    2006-01-01

    The author selected a simple random sample of 100 movies from the "Movie and Video Guide" (1996), by Leonard Maltin. The author's intent was to obtain some basic information on the population of roughly 19,000 movies through a small sample. The "Movie and Video Guide" by Leonard Maltin is an annual ratings guide to movies. While not all films ever…

  18. Introduction to Film.

    ERIC Educational Resources Information Center

    Burns, Gary

    There are numerous ways to structure the introduction to film course so as to meet the needs of the different types of students who typically enroll. Assuming there is no production component in the course, the teacher is left with two major approaches to choose from--historical and aesthetic. The units in the course will typically be built around…

  19. Thin film photovoltaic cell

    DOEpatents

    Meakin, John D.; Bragagnolo, Julio

    1982-01-01

    A thin film photovoltaic cell having a transparent electrical contact and an opaque electrical contact with a pair of semiconductors therebetween includes utilizing one of the electrical contacts as a substrate and wherein the inner surface thereof is modified by microroughening while being macro-planar.

  20. Diamond films: Historical perspective

    SciTech Connect

    Messier, R.

    1993-01-01

    This section is a compilation of notes and published international articles about the development of methods of depositing diamond films. Vapor deposition articles are included from American, Russian, and Japanese publications. The international competition to develop new deposition methodologies is stressed. The current status of chemical vapor deposition of diamond is assessed.

  1. Films on Deafness.

    ERIC Educational Resources Information Center

    Parlato, Salvatore J., Jr., Comp.

    This filmography on deafness, which contains summaries of 192 16mm films arranged in alphabetical order by title, covers a wide variety of topics as evidenced by the categorical title index: communication, the nature of deafness, detection and measurement of deafness, education and training, multi-handicaps, and noise pollution. Running time, date…

  2. Films in Depth.

    ERIC Educational Resources Information Center

    Schrievogel, Paul A.; Prete, Anthony T.

    Bound in a slipcover rather than in signatures, this "book" is made up of thirteen separately bound booklets. The first booklet is an introduction to the use of film in the classroom both in teaching the filmic art and in increasing the visual literacy of students on the high school and early college levels. The twelve other booklets each treat a…

  3. A Film Canister Colorimeter.

    ERIC Educational Resources Information Center

    Gordon, James; James, Alan; Harman, Stephanie; Weiss, Kristen

    2002-01-01

    A low-cost, low-tech colorimeter was constructed from a film canister. The student-constructed colorimeter was used to show the Beer-Lambert relationship between absorbance and concentration and to calculate the value of the molar absorptivity for permanganate at the wavelength emission maximum for an LED. Makes comparisons between this instrument…

  4. Rating Films on TV.

    ERIC Educational Resources Information Center

    Herman, Ginette; Leyens, Jacques-Philippe

    1977-01-01

    An analysis of the film viewing habits of Belgian television viewers reveals that movies with advisories regarding sex and violence are watched more than the movies without them. However, movies with qualifications tend to be judged less interesting than movies without qualifications. (JMF)

  5. Films: 1971/72 Catalog of Films, Film Loops and Filmstrips for Schools, Colleges and Libraries.

    ERIC Educational Resources Information Center

    Learning Corp. of America, New York, NY.

    The films in this catalog are available for sale or rent from Learning Corporation of America. For elementary grades, films are available for use in the language arts and social studies classes. For junior and senior high, college, and adult courses, films are listed for instruction in art, music, and dance; environmental studies; United States…

  6. Orientation filtering for crystalline films

    DOEpatents

    Smith, H.I.; Atwater, H.A.; Thompson, C.V.; Geis, M.W.

    1986-12-30

    A substrate is coated with a film to be recrystallized. A pattern of crystallization barriers is created in the film, for example, by etching voids in the film. An encapsulation layer is generally applied to protect the film, fill the voids and otherwise enhance a recrystallization process. Recrystallization is carried out such that certain orientations pass preferentially through the barrier, generally as a result of growth-velocity anisotropy. The result is a film of a specific predetermined crystallographic orientation, a range of orientations or a set of discrete orientations. 7 figs.

  7. Orientation filtering for crystalline films

    DOEpatents

    Smith, Henry I.; Atwater, Harry A.; Thompson, Carl V.; Geis, Michael W.

    1986-12-30

    A substrate is coated with a film to be recrystallized. A pattern of crystallization barriers is created in the film, for example, by etching voids in the film. An encapsulation layer is generally applied to protect the film, fill the voids and otherwise enhance a recrystallization process. Recrystallization is carried out such that certain orientations pass preferentially through the barrier, generally as a result of growth-velocity anisotropy. The result is a film of a specific predetermined crystallographic orientation, a range of orientations or a set of discrete orientations.

  8. Process to form mesostructured films

    DOEpatents

    Brinker, C. Jeffrey; Anderson, Mark T.; Ganguli, Rahul; Lu, Yunfeng

    1999-01-01

    This invention comprises a method to form a family of supported films film with pore size in the approximate range 0.8-20 nm exhibiting highly ordered microstructures and porosity derived from an ordered micellar or liquid-crystalline organic-inorganic precursor structure that forms during film deposition. Optically transparent, 100-500-nm thick films exhibiting a unique range of microstructures and uni-modal pore sizes are formed in seconds in a continuous coating operation. Applications of these films include sensors, membranes, low dielectric constant interlayers, anti-reflective coatings, and optical hosts.

  9. Why do drying films crack?

    PubMed

    Lee, Wai Peng; Routh, Alexander F

    2004-11-01

    Understanding the mechanism by which films fail during drying is the first step in controlling this natural process. Previous studies have examined the spacing between cracks with predictions made by assuming a balance between elastic energy released with a surface energy consumed. We introduce a new scaling for the spacing between cracks in drying dispersions. The scaling relates to the distance that solvent can flow, to relieve capillary stresses, as a film fails. The scaling collapses data for a range of evaporation rates, film thicknesses, particle sizes, and materials. This work identifies capillary pressures, induced by packed particle fronts travelling horizontally across films, as responsible for the failure in dried films. PMID:15518466

  10. Thin-film optical initiator

    DOEpatents

    Erickson, Kenneth L.

    2001-01-01

    A thin-film optical initiator having an inert, transparent substrate, a reactive thin film, which can be either an explosive or a pyrotechnic, and a reflective thin film. The resultant thin-film optical initiator system also comprises a fiber-optic cable connected to a low-energy laser source, an output charge, and an initiator housing. The reactive thin film, which may contain very thin embedded layers or be a co-deposit of a light-absorbing material such as carbon, absorbs the incident laser light, is volumetrically heated, and explodes against the output charge, imparting about 5 to 20 times more energy than in the incident laser pulse.

  11. Combustion effects on film cooling

    NASA Technical Reports Server (NTRS)

    Rousar, D. C.; Ewen, R. L.

    1977-01-01

    The effects of: (1) a reactive environment on film cooling effectiveness, and (2) film cooling on rocket engine performance were determined experimentally in a rocket thrust chamber assembly operating with hydrogen and oxygen propellants at 300 psi chamber pressure. Tests were conducted using hydrogen, helium, and nitrogen film coolants in an instrumented, thin walled, steel thrust chamber. The film cooling, performance loss, and heat transfer coefficient data were correlated with the ALRC entrainment film cooling model which relates film coolant effectiveness and mixture ratio at the wall to the amount of mainstream gases entrained with the film coolant in a mixing layer. In addition, a comprehensive thermal analysis computer program, HOCOOL, was prepared from previously existing ALRC computer programs and analytical techniques.

  12. Professor Camillo Negro's Neuropathological Films.

    PubMed

    Chiò, Adriano; Gianetto, Claudia; Dagna, Stella

    2016-01-01

    Camillo Negro, Professor in Neurology at the University of Torino, was a pioneer of scientific film. From 1906 to 1908, with the help of his assistant Giuseppe Roasenda and in collaboration with Roberto Omegna, one of the most experienced cinematographers in Italy, he filmed some of his patients for scientific and educational purposes. During the war years, he continued his scientific film project at the Military Hospital in Torino, filming shell-shocked soldiers. In autumn 2011, the Museo Nazionale del Cinema, in partnership with the Faculty of Neurosciences of the University of Torino, presented a new critical edition of the neuropathological films directed by Negro. The Museum's collection also includes 16 mm footage probably filmed in 1930 by Doctor Fedele Negro, Camillo's son. One of these films is devoted to celebrating the effects of the so-called "Bulgarian cure" on Parkinson's disease. PMID:26684422

  13. Thin film superconductor magnetic bearings

    DOEpatents

    Weinberger, Bernard R.

    1995-12-26

    A superconductor magnetic bearing includes a shaft (10) that is subject to a load (L) and rotatable around an axis of rotation, a magnet (12) mounted to the shaft, and a stator (14) in proximity to the shaft. The stator (14) has a superconductor thin film assembly (16) positioned to interact with the magnet (12) to produce a levitation force on the shaft (10) that supports the load (L). The thin film assembly (16) includes at least two superconductor thin films (18) and at least one substrate (20). Each thin film (18) is positioned on a substrate (20) and all the thin films are positioned such that an applied magnetic field from the magnet (12) passes through all the thin films. A similar bearing in which the thin film assembly (16) is mounted on the shaft (10) and the magnet (12) is part of the stator (14) also can be constructed.

  14. Method for making carbon films

    DOEpatents

    Tan, M.X.

    1999-07-29

    A method for treating an organic polymer material, preferably a vinylidene chloride/vinyl chloride copolymer (Saran) to produce a flat sheet of carbon film material having a high surface area ([approx equal]1000 m[sup 2] /g) suitable as an electrode material for super capacitor applications. The method comprises heating a vinylidene chloride/vinyl chloride copolymer film disposed between two spaced apart graphite or ceramic plates to a first temperature of about 160 C for about 14 hours to form a stabilized vinylidene chloride/vinyl chloride polymer film, thereafter heating the stabilized film to a second temperature of about 750 C in an inert atmosphere for about one hour to form a carbon film; and finally activating the carbon film to increase the surface area by heating the carbon film in an oxidizing atmosphere to a temperature of at least 750--850 C for between 1--6 hours. 2 figs.

  15. Method for making carbon films

    DOEpatents

    Tan, Ming X.

    1999-01-01

    A method for treating an organic polymer material, preferably a vinylidene chloride/vinyl chloride copolymer (Saran) to produce a flat sheet of carbon film material having a high surface area (.apprxeq.1000 m.sup.2 /g) suitable as an electrode material for super capacitor applications. The method comprises heating a vinylidene chloride/vinyl chloride copolymer film disposed between two spaced apart graphite or ceramic plates to a first temperature of about 160.degree. C. for about 14 hours to form a stabilized vinylidene chloride/vinyl chloride polymer film, thereafter heating the stabilized film to a second temperature of about 750.degree. C. in an inert atmosphere for about one hour to form a carbon film; and finally activating the carbon film to increase the surface area by heating the carbon film in an oxidizing atmosphere to a temperature of at least 750-850.degree. C. for between 1-6 hours.

  16. Thin film superconductor magnetic bearings

    SciTech Connect

    Weinberger, B.R.

    1995-12-26

    A superconductor magnetic bearing includes a shaft that is subject to a load (L) and rotatable around an axis of rotation, a magnet mounted to the shaft, and a stator in proximity to the shaft. The stator has a superconductor thin film assembly positioned to interact with the magnet to produce a levitation force on the shaft that supports the load (L). The thin film assembly includes at least two superconductor thin films and at least one substrate. Each thin film is positioned on a substrate and all the thin films are positioned such that an applied magnetic field from the magnet passes through all the thin films. A similar bearing in which the thin film assembly is mounted on the shaft and the magnet is part of the stator also can be constructed. 8 figs.

  17. Host thin films incorporating nanoparticles

    NASA Astrophysics Data System (ADS)

    Qureshi, Uzma

    The focus of this research project was the investigation of the functional properties of thin films that incorporate a secondary nanoparticulate phase. In particular to assess if the secondary nanoparticulate material enhanced a functional property of the coating on glass. In order to achieve this, new thin film deposition methods were developed, namely use of nanopowder precursors, an aerosol assisted transport technique and an aerosol into atmospheric pressure chemical vapour deposition system. Aerosol assisted chemical vapour deposition (AACVD) was used to deposit 8 series of thin films on glass. Five different nanoparticles silver, gold, ceria, tungsten oxide and zinc oxide were tested and shown to successfully deposit thin films incorporating nanoparticles within a host matrix. Silver nanoparticles were synthesised and doped within a titania film by AACVD. This improved solar control properties. A unique aerosol assisted chemical vapour deposition (AACVD) into atmospheric pressure chemical vapour deposition (APCVD) system was used to deposit films of Au nanoparticles and thin films of gold nanoparticles incorporated within a host titania matrix. Incorporation of high refractive index contrast metal oxide particles within a host film altered the film colour. The key goal was to test the potential of nanopowder forms and transfer the suspended nanopowder via an aerosol to a substrate in order to deposit a thin film. Discrete tungsten oxide nanoparticles or ceria nanoparticles within a titanium dioxide thin film enhanced the self-cleaning and photo-induced super-hydrophilicity. The nanopowder precursor study was extended by deposition of zinc oxide thin films incorporating Au nanoparticles and also ZnO films deposited from a ZnO nanopowder precursor. Incorporation of Au nanoparticles within a VO: host matrix improved the thermochromic response, optical and colour properties. Composite VC/TiC and Au nanoparticle/V02/Ti02 thin films displayed three useful

  18. Thin film mechanics

    NASA Astrophysics Data System (ADS)

    Cooper, Ryan C.

    This doctoral thesis details the methods of determining mechanical properties of two classes of novel thin films suspended two-dimensional crystals and electron beam irradiated microfilms of polydimethylsiloxane (PDMS). Thin films are used in a variety of surface coatings to alter the opto-electronic properties or increase the wear or corrosion resistance and are ideal for micro- and nanoelectromechanical system fabrication. One of the challenges in fabricating thin films is the introduction of strains which can arise due to application techniques, geometrical conformation, or other spurious conditions. Chapters 2-4 focus on two dimensional materials. This is the intrinsic limit of thin films-being constrained to one atomic or molecular unit of thickness. These materials have mechanical, electrical, and optical properties ideal for micro- and nanoelectromechanical systems with truly novel device functionality. As such, the breadth of applications that can benefit from a treatise on two dimensional film mechanics is reason enough for exploration. This study explores the anomylously high strength of two dimensional materials. Furthermore, this work also aims to bridge four main gaps in the understanding of material science: bridging the gap between ab initio calculations and finite element analysis, bridging the gap between ab initio calculations and experimental results, nanoscale to microscale, and microscale to mesoscale. A nonlinear elasticity model is used to determine the necessary elastic constants to define the strain-energy density function for finite strain. Then, ab initio calculations-density functional theory-is used to calculate the nonlinear elastic response. Chapter 2 focuses on validating this methodology with atomic force microscope nanoindentation on molybdenum disulfide. Chapter 3 explores the convergence criteria of three density functional theory solvers to further verify the numerical calculations. Chapter 4 then uses this model to investigate

  19. thin films as absorber

    NASA Astrophysics Data System (ADS)

    González, J. O.; Shaji, S.; Avellaneda, D.; Castillo, G. A.; Das Roy, T. K.; Krishnan, B.

    2014-09-01

    Photovoltaic structures were prepared using AgSb(S x Se1- x )2 as absorber and CdS as window layer at various conditions via a hybrid technique of chemical bath deposition and thermal evaporation followed by heat treatments. Silver antimony sulfo selenide thin films [AgSb(S x Se1- x )2] were prepared by heating multilayers of sequentially deposited Sb2S3/Ag dipped in Na2SeSO3 solution, glass/Sb2S3/Ag/Se. For this, Sb2S3 thin films were deposited from a chemical bath containing SbCl3 and Na2S2O3. Then, Ag thin films were thermally evaporated on glass/Sb2S3, followed by selenization by dipping in an acidic solution of Na2SeSO3. The duration of dipping was varied as 3, 4 and 5 h. Two different heat treatments, one at 350 °C for 20 min in vacuum followed by a post-heat treatment at 325 °C for 2 h in Ar, and the other at 350 °C for 1 h in Ar, were applied to the multilayers of different configurations. X-ray diffraction results showed the formation of AgSb(S x Se1- x )2 thin films as the primary phase and AgSb(S,Se)2 and Sb2S3 as secondary phases. Morphology and elemental detection were done by scanning electron microscopy and energy dispersive X-ray analysis. X-ray photoelectron spectroscopic studies showed the depthwise composition of the films. Optical properties were determined by UV-vis-IR transmittance and reflection spectral analysis. AgSb(S x Se1- x )2 formed at different conditions was incorporated in PV structures glass/FTO/CdS/AgSb(S x Se1- x )2/C/Ag. Chemically deposited post-annealed CdS thin films of various thicknesses were used as window layer. J- V characteristics of the cells were measured under dark and AM1.5 illumination. Analysis of the J- V characteristics resulted in the best solar cell parameters of V oc = 520 mV, J sc = 9.70 mA cm-2, FF = 0.50 and η = 2.7 %.

  20. SU-E-T-112: Dose Distribution of Praseodymium-142 Microspheres in Microcapillary Using Radiochromic Film Dosimetry and Applications in Hepatocellular Carcinoma Microsphere Brachytherapy

    SciTech Connect

    Ferreira, M; Rasmussen, K; Jung, J

    2014-06-01

    Purpose: This work verified simulations of beta-minus emitter Praseodymium-142 (Pr-142) for microsphere brachytherapy by performing absolute dose measurements for Pr 142 microspheres in a microcapillary as a simplified model for a single blood vessel for the treatment of Hepatocellular Carcinoma (HCC). Methods: Pr-142 microspheres (mass: 0.169g, average diameter: 29.7±3.9μm) were activated by thermal neutron activation at the University of Missouri Research Reactor. Experimental setup consisted of a microsphere solution (initial activity 36.6mCi in 0.1ml of sterile water) within a glass microcapillary (internal and external diameter: 305μm and 453μm, respectively) placed for 51h in a custom made Gammex Solid Water™ phantom. GAFCHROMIC™ EBT2 film calibrated with a 6MeV electron beam was used to access the dose fall-off of microspheres. The microcapillary was modeled in MCNPX2.6 in order to compare with experiments. Results: The radial dose fall-off on the transverse plane due to scatter and attenuation in the solid water phantom was analyzed using ImageJ for both film and MCNPX2.6 simulations. Isodose analysis showed close agreement among the methods used, i.e. measurements and simulations agree within 3.9% for doses below 1600cGy. Experimental and simulated doses obtained at 0.5 cm radially from the source were 1547cGy and 1610cGy respectively. Discrepancies for points close to the microcapillary surface were observed between MCNPX2.6 and measurements due to film saturation for high doses. Dose due to Pr-142 3.7% gamma emission was below the threshold of detection for the film. Conclusion: A detailed dosimetric study was performed for Pr-142 glass microspheres within a single microcapillary. MCNPX2.6 simulations were verified by means of direct measurement. Based on these results, Pr-142 appears to be a viable choice of radionuclide for treating HCC.

  1. Rupture of vertical soap films

    NASA Astrophysics Data System (ADS)

    Rio, Emmanuelle

    2014-11-01

    Soap films are ephemeral and fragile objects. They tend to thin under gravity, which gives rise to the fascinating variations of colors at their interfaces but leads systematically to rupture. Even a child can create, manipulate and admire soap films and bubbles. Nevertheless, the reason why it suddenly bursts remains a mystery although the soap chosen to stabilize the film as well as the humidity of the air seem very important. One difficulty to study the rupture of vertical soap films is to control the initial solution. To avoid this problem we choose to study the rupture during the generation of the film at a controlled velocity. We have built an experiment, in which we measure the maximum length of the film together with its lifetime. The generation of the film is due to the presence of a gradient of surface concentration of surfactants at the liquid/air interface. This leads to a Marangoni force directed toward the top of the film. The film is expected to burst only when its weight is not balanced anymore by this force. We will show that this leads to the surprising result that the thicker films have shorter lifetimes than the thinner ones. It is thus the ability of the interface to sustain a surface concentration gradient of surfactants which controls its stability.

  2. Advanced thin film thermocouples

    NASA Astrophysics Data System (ADS)

    Kreider, K. G.; Semancik, S.; Olson, C.

    1984-10-01

    The fabrication, materials characterization, and performance of thin film platinum rhodium thermocouples on gas turbine alloys was investigated. The materials chosen for the study were the turbine blade alloy systems MAR M200+Hf with NiCoCrAlY and FeCrAlY coatings, and vane alloy systems MAR M509 with FeCrAlY. Research was focussed on making improvements in the problem areas of coating substrate stability, adhesion, and insulation reliability and durability. Diffusion profiles between the substrate and coating with and without barrier coatings of Al2O3 are reported. The relationships between fabrication parameters of thermal oxidation and sputtering of the insulator and its characterization and performance are described. The best thin film thermocouples were fabricated with the NiCoCrAlY coatings which were thermally oxidized and sputter coated with Al2O3.

  3. Freely floating smectic films.

    PubMed

    May, Kathrin; Harth, Kirsten; Trittel, Torsten; Stannarius, Ralf

    2014-05-19

    We have investigated the dynamics of freely floating smectic bubbles using high-speed optical imaging. Bubbles in the size range from a few hundred micrometers to several centimeters were prepared from collapsing catenoids. They represent ideal model systems for the study of thin-film fluid dynamics under well-controlled conditions. Owing to the internal smectic layer structure, the bubbles combine features of both soap films and vesicles in their unique shape dynamics. From a strongly elongated initial shape after pinch-off, they relax towards the spherical equilibrium, first by a slow redistribution of the smectic layers, and finally by weak, damped shape oscillations. In addition, we describe the rupture of freely floating smectic bubbles, and the formation and stability of smectic filaments. PMID:24692347

  4. Biomimetic thin film deposition

    SciTech Connect

    Rieke, P.R.; Graff, G.E.; Campbell, A.A.; Bunker, B.C.; Baskaran, S.; Song, L.; Tarasevich, B.J.; Fryxell, G.E.

    1995-09-01

    Biological mineral deposition for the formation of bone, mollusk shell and other hard tissues provides materials scientists with illustrative materials processing strategies. This presentation will review the key features of biomineralization and how these features can be of technical importance. We have adapted existing knowledge of biomineralization to develop a unique method of depositing inorganic thin films and coating. Our approach to thin film deposition is to modify substrate surfaces to imitate the proteins found in nature that are responsible for controlling mineral deposition. These biomimetic surfaces control the nucleation and growth of the mineral from a supersaturated aqueous solution. This has many processing advantages including simple processing equipment, environmentally benign reagents, uniform coating of highly complex shapes, and enhanced adherence of coating. Many different types of metal oxide, hydroxide, sulfide and phosphate materials with useful mechanical, optical, electronic and biomedical properties can be deposited.

  5. Advanced thin film thermocouples

    NASA Technical Reports Server (NTRS)

    Kreider, K. G.; Semancik, S.; Olson, C.

    1984-01-01

    The fabrication, materials characterization, and performance of thin film platinum rhodium thermocouples on gas turbine alloys was investigated. The materials chosen for the study were the turbine blade alloy systems MAR M200+Hf with NiCoCrAlY and FeCrAlY coatings, and vane alloy systems MAR M509 with FeCrAlY. Research was focussed on making improvements in the problem areas of coating substrate stability, adhesion, and insulation reliability and durability. Diffusion profiles between the substrate and coating with and without barrier coatings of Al2O3 are reported. The relationships between fabrication parameters of thermal oxidation and sputtering of the insulator and its characterization and performance are described. The best thin film thermocouples were fabricated with the NiCoCrAlY coatings which were thermally oxidized and sputter coated with Al2O3.

  6. Molecular films associated with LDEF

    NASA Technical Reports Server (NTRS)

    Crutcher, E. R.; Warner, K. J.

    1992-01-01

    The molecular films deposited on the surface of the Long Duration Exposure Facility (LDEF) originated from the paints and room-temperature-vulcanized (RTV) silicone materials intentionally used on the satellite and not from residual contaminants. The high silicone content of most of the films and the uniformity of the films indicates a homogenization process in the molecular deposition and suggests a chemically most favored composition for the final film. The deposition on interior surfaces and vents indicated multiple bounce trajectories or repeated deposition-reemission cycles. Exterior surface deposits indicated a significant return flux. Ultraviolet light exposure was required to fix the deposited film as is indicated by the distribution of the films on interior surfaces and the thickness of films at the vent locations. Thermal conditions at the time of exposure to ultraviolet light seems to be an important factor in the thickness of the deposit. Sunrise facing (ram direction) surfaces always had the thicker film. These were the coldest surfaces at the time of their exposure to ultraviolet light. The films have a layered structure suggesting cyclic deposition. As many as 34 distinct layers were seen in the films. The cyclic nature of the deposition and the chemical uniformity of the film one layer to the next suggest an early deposition of the films though there is evidence for the deposition of molecular films throughout the nearly six year exposure of the satellite. A final 'spray' of an organic material associated with water soluble salts occurred very late in the mission. This may have been the result of one of the shuttle dump activities.

  7. Nonlinear optical thin films

    NASA Technical Reports Server (NTRS)

    Leslie, Thomas M.

    1993-01-01

    A focused approach to development and evaluation of organic polymer films for use in optoelectronics is presented. The issues and challenges that are addressed include: (1) material synthesis, purification, and the tailoring of the material properties; (2) deposition of uniform thin films by a variety of methods; (3) characterization of material physical properties (thermal, electrical, optical, and electro-optical); and (4) device fabrication and testing. Photonic materials, devices, and systems were identified as critical technology areas by the Department of Commerce and the Department of Defense. This approach offers strong integration of basic material issues through engineering applications by the development of materials that can be exploited as the active unit in a variety of polymeric thin film devices. Improved materials were developed with unprecedented purity and stability. The absorptive properties can be tailored and controlled to provide significant improvement in propagation losses and nonlinear performance. Furthermore, the materials were incorporated into polymers that are highly compatible with fabrication and patterning processes for integrated optical devices and circuits. By simultaneously addressing the issues of materials development and characterization, keeping device design and fabrication in mind, many obstacles were overcome for implementation of these polymeric materials and devices into systems. We intend to considerably improve the upper use temperature, poling stability, and compatibility with silicon based devices. The principal device application that was targeted is a linear electro-optic modulation etalon. Organic polymers need to be properly designed and coupled with existing integrated circuit technology to create new photonic devices for optical communication, image processing, other laser applications such as harmonic generation, and eventually optical computing. The progression from microscopic sample to a suitable film

  8. Falling film evaporator

    DOEpatents

    Bruns, Lester E.

    1976-01-01

    A falling film evaporator including a vertically oriented pipe heated exteriorly by a steam jacket and interiorly by a finned steam tube, all heating surfaces of the pipe and steam tube being formed of a material wet by water such as stainless steel, and packing within the pipe consisting of Raschig rings formed of a material that is not wet by water such as polyvinylidene fluoride.

  9. Thick film ink chemistry

    NASA Astrophysics Data System (ADS)

    Gehman, R. W.

    1982-03-01

    Twenty-six thick film inks from two vendors were proved for hybrid microcircuit production use. A data base of chemical information was established for all the inks to aid in future diagnostic and failure analysis activities. Efforts included both organic chemical analysis of printing vehicles and binders and inorganic chemical analysis of glass frits and electrically active phases. Analytical methods included infrared spectroscopy, mass spectroscopy, gas chromatography, X-ray fluorescence, emission spectroscopy, atomic absorption spectroscopy, and wet chemical techniques.

  10. Polycrystalline thin film photovoltaics

    NASA Astrophysics Data System (ADS)

    Zweibel, K.; Ullal, H. S.; Mitchell, R. L.

    Significant progress has recently been made towards improving the efficiencies of polycrystalline thin-film solar cells and modules using CuInSe2 and CdTe. The history of using CuInSe2 and CdTe for solar cells is reviewed. Initial outdoor stability tests of modules are encouraging. Progress in semiconductor deposition techniques has also been substantial. Both CuInSe2 and CdTe are positioned for commercialization during the 1990s. The major participants in developing these materials are described. The US DOE/SERI (Solar Energy Research Institute) program recognizes the rapid progress and important potential of polycrystalline thin films to meet ambitious cost and performance goals. US DOE/SERI is in the process of funding an initiative in this area with the goal of ensuring US leadership in the development of these technologies. The polycrystalline thin-film module development initiative, the modeling and stability of the devices, and health and safety issues are discussed.

  11. Spring magnet films.

    SciTech Connect

    Bader, S. D.; Fullerton, E. E.; Gornakov, V. S.; Inomata, A.; Jiang, J. S.; Nikitenko, V. I.; Shapiro, A. J.; Shull, R. D.; Sowers, C. H.

    1999-03-29

    The properties of exchange-spring-coupled bilayer and superlattice films are highlighted for Sm-Co hard magnet and Fe or Co soft magnet layers. The hexagonal Sm-Co is grown via magnetron sputtering in a- and b-axis epitaxial orientations. In both cases the c-axis, in the film plane, is the easy axis of magnetization. Trends in coercivity with film thickness are established and related to the respective microstructure of the two orientations. The magnetization reversal process for the bilayers is examined by magnetometry and magneto-optical imaging, as well as by simulations that utilize a one-dimensional model to provide the spin configuration for each atomic layer. The Fe magnetization is pinned to that of the Sm-Co at the interface, and reversal proceeds via a progressive twisting of the Fe magnetization. The Fe demagnetization curves are reversible as expected for a spring magnet. Comparison of experiment and simulations indicates that the spring magnet behavior can be understood from the intrinsic properties of the hard and soft layers. Estimated are made of the ultimate gain in performance that can potentially be realized in this system.

  12. Evaluation of superficial dosimetry between treatment planning system and measurement for several breast cancer treatment techniques

    SciTech Connect

    Akino, Yuichi; Das, Indra J.; Bartlett, Gregory K.; Zhang Hualin; Thompson, Elizabeth; Zook, Jennifer E.

    2013-01-15

    Purpose: Dosimetric accuracy in radiation treatment of breast cancer is critical for the evaluation of cosmetic outcomes and survival. It is often considered that treatment planning systems (TPS) may not be able to provide accurate dosimetry in the buildup region. This was investigated in various treatment techniques such as tangential wedges, field-in-field (FF), electronic compensator (eComp), and intensity-modulated radiotherapy (IMRT). Methods: Under Institutional Review Board (IRB) exemption, radiotherapy treatment plans of 111 cases were retrospectively analyzed. The distance between skin surface and 95% isodose line was measured. For measurements, Gafchromic EBT2 films were used on a humanoid unsliced phantom. Multiple layers of variable thickness of superflab bolus were placed on the breast phantom and CT scanned for planning. Treatment plans were generated using four techniques with two different grid sizes (1 Multiplication-Sign 1 and 2.5 Multiplication-Sign 2.5 mm{sup 2}) to provide optimum dose distribution. Films were placed at different depths and exposed with the selected techniques. A calibration curve for dose versus pixel values was also generated on the same day as the phantom measurement was conducted. The DICOM RT image, dose, and plan data were imported to the in-house software. On axial plane of CT slices, curves were drawn at the position where EBT2 films were placed, and the dose profiles on the lines were acquired. The calculated and measured dose profiles were separated by check points which were marked on the films before irradiation. The segments of calculated profiles were stretched to match their resolutions to that of film dosimetry. Results: On review of treatment plans, the distance between skin and 95% prescribed dose was up to 8 mm for plans of 27 patients. The film measurement revealed that the medial region of phantom surface received a mere 45%-50% of prescribed dose. For wedges, FF, and eComp techniques, region around the

  13. Evaluation of a commercially-available block for spatially fractionated radiation therapy.

    PubMed

    Buckey, Courtney; Stathakis, Sotirios; Cashon, Ken; Gutierrez, Alonso; Esquivel, Carlos; Shi, Chengyu; Papanikolaou, Nikos

    2010-01-01

    In this paper, we present the dosimetric characteristics of a commercially-produced universal GRID block for spatially fractioned radiation therapy. The dosimetric properties of the GRID block were evaluated. Ionization chamber and film measurements using both Kodak EDR2 and Gafchromic EBT film were performed in a solid water phantom to determine the relative output of the GRID block as well as its spatial dosimetric characteristics. The surface dose under the block and at the openings was measured using ultra thin TLDs. After introducing the GRID block into the treatment planning system, a treatment plan was created using the GRID block and also by creating a GRID pattern using the multi-leaf collimator. The percent depth doses measured with film showed that there is a shift of the dmax towards shallower depths for both energies (6 MV and 18 MV) under investigation. It was observed that the skin dose at the GRID openings was higher than the corresponding open field by a factor as high as 50% for both photon energies. The profiles showed the transmission under the block was in the order of 15-20% for 6 MV and 30% for 18 MV. The MUs calculated for a real patient using the block were about 80% less than the corresponding MUs for the same plan using the multileaf collimator to define the GRID. Based on this investigation, this brass GRID compensator is a viable alternative to other solid compensators or MLC-based fields currently in use. Its ease of creation and use give it decided advantages. Its ability to be created once and used for multiple patients (by varying the collimation of the linear accelerator jaws) makes it attractive from a cost perspective. We believe this compensator can be put to clinical use, and will allow more centers to offer GRID therapy to their patients. PMID:20717082

  14. SU-E-T-604: Penumbra Characteristics of a New InCiseâ„¢ Multileaf Collimator of CyberKnife M6â„¢ System

    SciTech Connect

    Hwang, M; Jang, S; Ozhasoglu, C; Lalonde, R; Heron, D; Huq, M

    2015-06-15

    Purpose: The InCise™ Multileaf Collimator (MLC) of CyberKnife M6™ System has been released recently. The purpose of this study was to explore the dosimetric characteristics of the new MLC. In particular, the penumbra characteristics of MLC fields at varying locations are evaluated. Methods: EBT3-based film measurements were performed with varying MLC fields ranging from 7.5 mm to 27.5 mm. Seventeen regions of interests (ROIs) were identified for irradiation. These are regions located at the central area (denoted as reference field), at the left/right edge areas of reference open field, at an intermediate location between central and edge area. Single beam treatment plans were designed by using the MultiPlan and was delivered using the Blue Phantom. Gafchromic films were irradiated at 1.5 cm depth in the Blue Phantom and analyzed using the Film Pro software. Variation of maximum dose, penumbra of MLC-defined fields, and symmetry/flatness were calculated as a function of locations of MLC fields. Results: The InCise™ MLC System showed relatively consistent dose distribution and penumbra size with varying locations of MLC fields. The measured maximum dose varied within 5 % at different locations compared to that at the central location and agreed with the calculated data well within 2%. The measured penumbrae were in the range of 2.9 mm and 3.7 mm and were relatively consistent regardless of locations. However, dose profiles in the out-of-field and in-field regions varied with locations and field sizes. Strong variation was seen for all fields located at 55 mm away from the central field. The MLC leakage map showed that the leakage is dependent on position. Conclusion: The size of penumbra and normalized maximum dose for MLC-defined fields were consistent in different regions of MLC. However, dose profiles in the out-field region varied with locations and field sizes.

  15. Determination of surface dose rate of indigenous (32)P patch brachytherapy source by experimental and Monte Carlo methods.

    PubMed

    Kumar, Sudhir; Srinivasan, P; Sharma, S D; Saxena, Sanjay Kumar; Bakshi, A K; Dash, Ashutosh; Babu, D A R; Sharma, D N

    2015-09-01

    Isotope production and Application Division of Bhabha Atomic Research Center developed (32)P patch sources for treatment of superficial tumors. Surface dose rate of a newly developed (32)P patch source of nominal diameter 25 mm was measured experimentally using standard extrapolation ionization chamber and Gafchromic EBT film. Monte Carlo model of the (32)P patch source along with the extrapolation chamber was also developed to estimate the surface dose rates from these sources. The surface dose rates to tissue (cGy/min) measured using extrapolation chamber and radiochromic films are 82.03±4.18 (k=2) and 79.13±2.53 (k=2) respectively. The two values of the surface dose rates measured using the two independent experimental methods are in good agreement to each other within a variation of 3.5%. The surface dose rate to tissue (cGy/min) estimated using the MCNP Monte Carlo code works out to be 77.78±1.16 (k=2). The maximum deviation between the surface dose rates to tissue obtained by Monte Carlo and the extrapolation chamber method is 5.2% whereas the difference between the surface dose rates obtained by radiochromic film measurement and the Monte Carlo simulation is 1.7%. The three values of the surface dose rates of the (32)P patch source obtained by three independent methods are in good agreement to one another within the uncertainties associated with their measurements and calculation. This work has demonstrated that MCNP based electron transport simulations are accurate enough for determining the dosimetry parameters of the indigenously developed (32)P patch sources for contact brachytherapy applications. PMID:26086681

  16. SU-E-CAMPUS-J-01: TG142 Complied Comprehensive Commissioning and Quality Assurance Procedure for Respiratory Gating

    SciTech Connect

    Woods, K; Rong, Y; Weldon, M; Gupta, N

    2014-06-15

    Purpose: To develop and establish a comprehensive gating commissioning and quality assurance procedure in compliance with TG142. Methods: Quality assurance tests on three Varian LINACs included beam output and energy constancy, calibration of surrogate, as well as phase and amplitude gating temporal accuracy. A diode array (MapCHECK 2) and film (Gafchromic EBT2) were used to measure the temporal accuracy of phase and amplitude gating windows. A motion simulation device (MotionSim) was used to simulate respiratory motion for both detectors. An end-to-end test was also performed on all three machines. The overall accuracy and uncertainty was analyzed and compared. Results: The end-to-end test using an anthropomorphic lung phantom (CIRS) results in an OSL dose difference reading within 5% (within measurement uncertainty) for both phase and amplitude gated treatment. Film results showed < 1% agreement between profiles for gated delivery and predicted dose. The diode array demonstrated an 80% passing rate for gamma criteria of 2%/0.2 mm, which results in a 111 msec temporal accuracy. However, the diode array is limited by its spatial resolution of measurements, due to its 7.07 mm diode spacing. Film provided higher measuring resolution, thus demonstrated a temporal accuracy of <100 msec. Conclusion: Results showed consistent respiratory gating stability and accuracy. MapCHECK 2 may not be sufficient for the temporal accuracy test in the respiratory gating treatment in order to meet the corresponding tolerance in TG142. One would need to decrease respiratory motion speed from the surrogate or tighten the gating window in order to be within tolerance of 100 msec temporal accuracy per TG-142. The end-to-end test offers insight to the overall accuracy and uncertainties with a gated protocol. Compared to static delivery, respiratory motion increases the overall uncertainty of treatment delivery from 3% to 5% dose difference.

  17. Evaluation of the dosimetric accuracy for a couch-based tracking system (CBTS)

    NASA Astrophysics Data System (ADS)

    Chang, Kyung Hwan; Lee, Suk; Kim, Kwang Hyeon; Shim, Jang Bo; Yang, Dae Sik; Park, Young Je; Yoon, Won Sup; Kim, Chul Yong; Cao, Yuanjie

    2016-07-01

    In this study, the geometric and dosimetric accuracy of an in-house-developed couch-based tracking system (CBTS) was investigated using both film and in-house-developed polymer gel dosimeters. We evaluated the 1D and the 2D motion accuracies of our couch system by using Gafchromic EBT film. For the 1D test, the couch system was moved 5, 10, and 20 mm in the X, Y, and Z directions, respectively. Meanwhile, for the 2D test, it was moved along the XY, YZ, and ZX directions. We compared the profiles, full widths at half maximum (FWHMs), and penumbras between the static and the tracking fields. For the 3D test, we quantitatively compared the dose distribution between the static and the tracking fields by using the polymer gel dosimeter when it was simultaneously moved in the XYZ directions. We confirmed that the film was moved according to motion amplitudes of 5, 10, and 20 mm in the X, Y, and Z directions, respectively, in the 1D and 2D motion tests. The value of the FWHM of the static field and the three tracking fields were 51.88, 53.28, 57.67, and 64.43 mm, respectively. Two types of penumbras became wider with increasing amplitudes compared to the static field. For the 3D test, the dose distribution of the XYZ tracking field was qualitatively larger than that of the static field. We conclude that this CBTS has the potential for pre-clinical applications in adaptive radiation therapy.

  18. Short Films for Physics Teaching, A Catalog.

    ERIC Educational Resources Information Center

    Bluestone, Barbara Z.; Roth, Richard F.

    This annotated film catalog is a product of the Conference on Single Concept Films in College Physics Teaching sponsored by the Commission on College Physics. Both 8mm and 16mm single concept films are listed for physics and related disciplines. The catalog includes commercial, noncommercial, and foreign films. However, the film coverage was…

  19. Films on the arms race

    SciTech Connect

    Dowling, J.

    1983-01-01

    Films convey the historical perspectives, the biographical stories, the insights of the participants, and the horror of nuclear war - far better than can any physicist. While films are not very efficient for covering details, derivation, or numbers, they can not be beaten in showing what really happens in a nuclear explosion, in getting across general concepts, in illustrating the parameters of a problem, and the problem itself. Most importantly, films and TV can reach the people who must be informed about these issues if we are to resolve the problems. The author points out how films can contribute to an understanding of the issues of the arms race and nuclear war, with references to specific films. An annotated bibliography of 37 films is then presented.

  20. Fracture characteristics of balloon films

    NASA Technical Reports Server (NTRS)

    Portanova, Marc A.

    1989-01-01

    An attempt was made to determine the failure modes of high altitude scientific balloons through an investigation of the fracture characteristics of the thin polyethylene films. Two films were the subject of the evaluation, Winzen Int.'s Stratafilm SF-85 and Raven Industries' Astro-E. Research began with an investigation of the film's cold brittleness point and it's effect on the ultimate strength and elasticity of the polyethylene film. A series of preliminary investigations were conducted to develop an understanding of the material characteristics. The primary focus of this investigation was on the notch sensitivity of the films. Simple stress strain tests were also conducted to enable analysis employing fracture toughness parameters. Studies were conducted on both film types at 23 C (room temperature), -60 C, -90 C, and -120 C.

  1. How to Read a Film: The Art, Technology, Language, History and Theory of Film and Media.

    ERIC Educational Resources Information Center

    Monaco, James

    This book discusses film as a narrative technique directly comparable to expression in prose narrative, in painting, and in music; it presents an overview of film as technology, the language of film and television, the history of film in America, Europe, and Asia, and the growth of film criticism. Chapters include "Film As an Art,""Technology:…

  2. Frequency mixer having ferromagnetic film

    DOEpatents

    Khitun, Alexander; Roshchin, Igor V.; Galatsis, Kosmas; Bao, Mingqiang; Wang, Kang L.

    2016-03-29

    A frequency conversion device, which may include a radiofrequency (RF) mixer device, includes a substrate and a ferromagnetic film disposed over a surface of the substrate. An insulator is disposed over the ferromagnetic film and at least one microstrip antenna is disposed over the insulator. The ferromagnetic film provides a non-linear response to the frequency conversion device. The frequency conversion device may be used for signal mixing and amplification. The frequency conversion device may also be used in data encryption applications.

  3. Diamond films for laser hardening

    NASA Technical Reports Server (NTRS)

    Albin, S.; Watkins, L.; Ravi, K.; Yokota, S.

    1989-01-01

    Laser-damage experiments were performed on free-standing polycrystalline diamond films prepared by plasma-enhanced CVD. The high laser-induced stress resistance found for this material makes it useful for thin-film coatings for laser optics. Results for diamond-coated silicon substrates demonstrate the enhanced damage threshold imparted by diamond thin-film coatings to materials susceptible to laser damage.

  4. Polycrystalline thin film photovoltaic technology

    SciTech Connect

    Ullal, H.S.; Zweibel, K.; Mitchell, R.L.; Noufi, R.

    1991-03-01

    Low-cost, high-efficiency thin-film modules are an exciting photovoltaic technology option for generating cost-effective electricity in 1995 and beyond. In this paper we review the significant technical progress made in the following thin films: copper indium diselenide, cadmium telluride, and polycrystalline thin silicon films. Also, the recent US DOE/SERI initiative to commercialize these emerging technologies is discussed. 6 refs., 9 figs.

  5. Chitosan composite films. Biomedical applications.

    PubMed

    Cárdenas, Galo; Anaya, Paola; von Plessing, Carlos; Rojas, Carlos; Sepúlveda, Jackeline

    2008-06-01

    Chitosan acetate films have been prepared using chitosans from shrimps (Pleuroncodes monodon) of low and high molecular weight (LMv = 68,000 g/mol and HMv = 232,000 g/mol) and deacetylation degree of 80 and 100%, respectively. The chitosan films were obtained by addition of several additives to acetic acid chitosan solutions, such as: glycerol, oleic acid and linoleic acid in different proportions. The pH of the solutions before casting ranged from 5.0 to 6.0. The composite film thickness are reported. The films have been analyzed by FTIR showing characteristic bands corresponding to the additives. The scanning electron microscopy (SEM) studies reveals the different morphology of the composite films. The films exhibit different physical properties depending upon the additives and/or mixture of them. The addition of glycerol to composite improves the elasticity of the films. The swelling in glucose and saline solutions for several films was evaluated, being higher in the glucose solution. The bactericide test against Staphylococcus aureus, Pseudomona aeruginosa and Acinetobacter baumanii in plates with either blood and or agar tripticase showed that the molecular weight influences on the bactericidal properties of the chitosan composite films and over its effect against gram positive and gram negative bacteria. Medical applications of the composite films were done in patients with burns, ulcers and injuries, the films containing glycerol showed good adhesion in comparison with those without it. The composite films tested were mainly three (1) chitosan acetate with glycerol, (2) chitosan acetate with oleic acid and (3) chitosan acetate with glycerol and oleic acid. Excellent results in the skin recovery were obtained after 7-10 days. Since the chitosan is biodegradable by the body enzymes it does not need to be removed and increases the gradual grows of the damage tissues. PMID:18165888

  6. Communicating Science: Lessons from Film.

    PubMed

    Berlin, Heather A

    2016-04-01

    Films engage us visually, aurally, viscerally, and emotionally. Incorporating science themes into films has the potential to open up new audiences to scientific ideas, pique their interests, and inspire them to engage in a broader discussion of the science itself. Here, I discuss several narrative techniques and strategies employed in film to effectively engage the audience around science themes, which may be useful tools for scientists looking to become better communicators. PMID:26988314

  7. Center for thin film studies

    NASA Astrophysics Data System (ADS)

    Shannon, Robert P.; Gibson, Ursula J.

    1987-11-01

    This report covers the first year of operation of the URI Thin Film Center (TFC), and describes a diverse array of studies on thin-film materials, substrates, and their processing and analysis. Individual efforts are highlighted in sections on nucleation studies, ion-assisted deposition, Rutherford backscattering spectrometry, Brillouin scattering, a continuum theory of the evolution of structure in thin films, a study of polishing parameters relevant to the preparation of substrates, and the setup of a characterization facility for the Center.

  8. Process to form mesostructured films

    DOEpatents

    Brinker, C.J.; Anderson, M.T.; Ganguli, R.; Lu, Y.F.

    1999-01-12

    This invention comprises a method to form a family of supported films with pore size in the approximate range 0.8-20 nm exhibiting highly ordered microstructures and porosity derived from an ordered micellar or liquid-crystalline organic-inorganic precursor structure that forms during film deposition. Optically transparent, 100-500-nm thick films exhibiting a unique range of microstructures and uni-modal pore sizes are formed in seconds in a continuous coating operation. Applications of these films include sensors, membranes, low dielectric constant interlayers, anti-reflective coatings, and optical hosts. 12 figs.

  9. Ratings for emotion film clips.

    PubMed

    Gabert-Quillen, Crystal A; Bartolini, Ellen E; Abravanel, Benjamin T; Sanislow, Charles A

    2015-09-01

    Film clips are widely utilized to elicit emotion in a variety of research studies. Normative ratings for scenes selected for these purposes support the idea that selected clips correspond to the intended target emotion, but studies reporting normative ratings are limited. Using an ethnically diverse sample of college undergraduates, selected clips were rated for intensity, discreteness, valence, and arousal. Variables hypothesized to affect the perception of stimuli (i.e., gender, race-ethnicity, and familiarity) were also examined. Our analyses generally indicated that males reacted strongly to positively valenced film clips, whereas females reacted more strongly to negatively valenced film clips. Caucasian participants tended to react more strongly to the film clips, and we found some variation by race-ethnicity across target emotions. Finally, familiarity with the films tended to produce higher ratings for positively valenced film clips, and lower ratings for negatively valenced film clips. These findings provide normative ratings for a useful set of film clips for the study of emotion, and they underscore factors to be considered in research that utilizes scenes from film for emotion elicitation. PMID:24984981

  10. Thin film of biocompatible polysaccharides

    NASA Astrophysics Data System (ADS)

    Richert, Ludovic; Lavalle, Philippe; Schaaf, Pierre; Voegel, Jean-Claude; Picart, Catherine

    2003-03-01

    The layer-by-layer deposition method proposed by Decher et al. (1991) is a very simple and versatile method used to build thin films. These films are of interest for bioengineering because of their unique properties and of the possible insertion of bioactive molecules. We present here the peculiar properties of a new kind of film formed with natural biopolymers, namely hyaluronan (HA)and chitosan (CHI). The films may be used as biomimetic substrates to control bacterial and cell adhesion. These polysaccharides are of particular interest because they are biodegradable, non toxic, and can be found in various tissues. Hyaluronan is also a natural ligand for a numerous type of cells through the CD44 receptor. Chitosan has already largely been used for its biological and anti-microbial properties. (CHI/HA) films were built in acidic pH at different ionic strength. The buildup was followed in situ by optical waveguide lightmode spectroscopy (OWLS), quartz crystal microbalance, streaming potential measurements and atomic force microscopy. The kinetics of adsorption and desorption of the polyelectrolytes depended on the ionic strength. Small islands were initially present on the surface which grew by mutual coalescence until becoming a flat film. The films were around 200 nm in thickness. These results suggest that different types of thin films constituted of polysaccharides can be built on any type of surface. These films are currently investigated toward their cell adhesion and bacterial adhesion properties.

  11. Optical limiting by chemically enhanced bacteriorhodopsin films

    NASA Astrophysics Data System (ADS)

    Song, Q. Wang; Zhang, Chungping; Gross, Richard; Birge, Robert

    1993-05-01

    Measurements of effective nonlinearity of a chemically enhanced bacteriorhodopsin film are presented, using 2-scan method. Optical limiting properties and the film's nonlinear transmission properties of the film are also studied.

  12. Film Review. Films for Use in Professional Courses.

    ERIC Educational Resources Information Center

    Graham, Mike, Ed.

    This book contains a detailed review of over ninety films in the professional preparation of teachers. It is one of eight publications which have been produced by the Science Teacher Education Project. Some of the films have become integral parts of the learning experiences which college and university people have prepared and others are…

  13. The Environment Film Review. A Critical Guide to Ecology Films.

    ERIC Educational Resources Information Center

    Environment Information Center, New York, NY.

    Critical reviews of more than 600 environmental films selected from a field of several thousand are contained in this reference guide. Designed to provide a comprehensive selection of films covering all aspects of environmental affairs, from air pollution to wildlife, the guide is primarily user-oriented. It consists of two parts: a Review…

  14. "Kuleshov on Film": A Spectator-Centered Film Theory.

    ERIC Educational Resources Information Center

    Curran, Trisha

    This paper describes some of the theories of cinematography of Soviet film theorist and filmmaker Lev Kuleshov. It points out that for him, film was communication portraying people's activities emanating from the environment. It explains that he was especially interested in audience response, particularly that of the proletariat, and that he felt…

  15. Radial mode structure of curvature-driven instabilities in EBT

    SciTech Connect

    Spong, D.A.

    1983-01-11

    Viewgraphs describe the theoretical treatment of the radial mode structure of plasma instabilities in the Elmo Bumpy Torus. The calculation retains nonlocal structure of the modes, connects inner and outer ring regions together, uses a self-consistent finite ..beta.., includes the relativistic effects for the hot electron ring, and examines a wide range of parameters. (WRF)

  16. Protective Film Moves Aside

    NASA Technical Reports Server (NTRS)

    2008-01-01

    Relatively warmer daytime temperatures on Mars have allowed the biobarrier -- a shiny, protective film -- to peel away a little more from the robotic arm of NASA's Phoenix Mars Lander.

    This image shows the spacecraft's robotic arm in its stowed configuration, with the biobarrier unpeeled on landing day, or Sol (Martian day) 0, and the lander's first full day on Mars, Sol 1.

    The 'elbow' of the arm can be seen at the top center of the picture, and the biobarrier is the shiny film seen to the left of the arm.

    The biobarrier is an extra precaution to protect Mars from contamination with any bacteria from Earth. While the whole spacecraft was decontaminated through cleaning, filters and heat, the robotic arm was given additional protection because it is the only spacecraft part that will directly touch the ice below the surface of Mars.

    Before the arm was heated, it was sealed in the biobarrier, which is made of a trademarked film called Tedlar that holds up to baking like a turkey-basting bag. This ensures that any new bacterial spores that might have appeared during the final steps before launch and during the journey to Mars will not contact the robotic arm.

    After Phoenix landed, springs were used to pop back the barrier, giving it room to deploy.

    These images were taken on May 25, 2008 and May 26, 2008 by the spacecraft's Surface Stereo Imager.

    The Phoenix Mission is led by the University of Arizona, Tucson, on behalf of NASA. Project management of the mission is by NASA's Jet Propulsion Laboratory, Pasadena, Calif. Spacecraft development is by Lockheed Martin Space Systems, Denver.

  17. Antibacterial Biomimetic Hybrid Films

    PubMed Central

    Ferrer, M. Carme Coll; Hickok, Noreen J.; Eckmann, David M.; Composto, Russell J.

    2012-01-01

    In this work, we present a novel method to prepare a hybrid coating based on dextran grafted to a substrate and embedded with silver nanoparticles (Ag NPs). First, the Ag NPs are synthesized in situ in the presence of oxidized dextran in solution. Second, the oxidized dextran is exposed to an amine functionalized surface resulting in the simultaneous grafting of dextran and the trapping of Ag NPs within the layer. The NP loading is controlled by the concentration of silver nitrate, which is 2 mM (DEX-Ag2) and 5 mM (DEX-Ag5). The dried film thickness increases with silver nitrate concentration from 2 nm for dextran to 7 nm and 12 nm for DEX-Ag2 and DEX-Ag5, respectively. The grafted dextran film displays features with a diameter and height of ~ 50 nm and 2 nm, respectively. For the DEX-Ag2 and DEX-Ag5, the dextran features as well as individual Ag NPs (~ 5 nm) and aggregates of Ag NPs are observed. Larger and more irregular aggregates are observed for DEX-Ag5. Overall, the Ag NPs are embedded in the dextran film as suggested by AFM and UVO studies. In terms of its antimicrobial activity, DEX-Ag2 resists bacterial adhesion to a greater extent than DEX-Ag5, which in turn is better than dextran and silicon. Because these antibacterial hybrid coatings can be grafted to a variety of surfaces, many biomedical applications can be envisioned, ranging from coating implants to catheters. PMID:23807896

  18. Radiation grafting on natural films

    NASA Astrophysics Data System (ADS)

    Lacroix, M.; Khan, R.; Senna, M.; Sharmin, N.; Salmieri, S.; Safrany, A.

    2014-01-01

    Different methods of polymer grafting using gamma irradiation are reported in the present study for the preparation of newly functionalized biodegradable films, and some important properties related to their mechanical and barrier properties are described. Biodegradable films composed of zein and poly(vinyl alcohol) (PVA) were gamma-irradiated in presence of different ratios of acrylic acid (AAc) monomer for compatibilization purpose. Resulting grafted films (zein/PVA-g-AAc) had their puncture strength (PS=37-40 N mm-1) and puncture deformation (PD=6.5-9.8 mm) improved for 30% and 50% PVA in blend, with 5% AAc under 20 kGy. Methylcellulose (MC)-based films were irradiated in the presence of 2-hydroxyethyl methacrylate (HEMA) or silane, in order to determine the effect of monomer grafting on the mechanical properties of films. It was found that grafted films (MC-g-HEMA and MC-g-silane) using 35% monomer performed higher mechanical properties with PS values of 282-296 N mm-1 and PD of 5.0-5.5 mm under 10 kGy. Compatibilized polycaprolactone (PCL)/chitosan composites were developed via grafting silane in chitosan films. Resulting trilayer grafted composite film (PCL/chitosan-g-silane/PCL) presented superior tensile strength (TS=22 MPa) via possible improvement of interfacial adhesion (PCL/chitosan) when using 25% silane under 10 kGy. Finally, MC-based films containing crystalline nanocellulose (CNC) as a filling agent were prepared and irradiated in presence of trimethylolpropane trimethacrylate (TMPTMA) as a grafted plasticizer. Grafted films (MC-g-TMPTMA) presented superior mechanical properties with a TS of 47.9 MPa and a tensile modulus (TM) of 1792 MPa, possibly due to high yield formation of radicals to promote TMPTMA grafting during irradiation. The addition of CNC led to an additional improvement of the barrier properties, with a significant 25% reduction of water vapor permeability (WVP) of grafted films.

  19. Electrically conductive palladium-containing polyimide films

    NASA Technical Reports Server (NTRS)

    St.clair, A. K.; Furtsch, T. A.; Taylor, L. T.

    1981-01-01

    Palladium addition makes light, flexible film with low resistivity to relieve space charging. Polyimide film is prepared in four steps: preparation of polyamic acid in polar solvent; addition of soluable palladium complex salt; fabrication of film of "palladium polyamic acid" solution; and thermal imidization of film to palladium-containing polyimide by 300 C heating. Lowered resistivities were achieved without loss in film flexibility or increase in film weight.

  20. Turbine airfoil film cooling

    NASA Astrophysics Data System (ADS)

    Hylton, Larry D.

    1986-10-01

    Emphasis is placed on developing more accurate analytical models for predicting turbine airfoil external heat transfer rates. Performance goals of new engines require highly refined, accurate design tools to meet durability requirements. In order to obtain improvements in analytical capabilities, programs are required which focus on enhancing analytical techniques through verification of new models by comparison with relevant experimental data. The objectives of the current program are to develop an analytical approach, based on boundary layer theory, for predicting the effects of airfoil film cooling on downstream heat transfer rates and to verify the resulting analytical method by comparison of predictions with hot cascade data obtained under this program.