Sample records for gafchromic ebt film

  1. Comparing three UV wavelengths for pre‐exposing Gafchromic EBT2 and EBT3 films

    PubMed Central

    Gotanda, Rumi; Gotanda, Tatsuhiro; Akagawa, Takuya; Tanki, Nobuyoshi; Kuwano, Tadao; Yabunaka, Kouichi

    2015-01-01

    Gafchromic films are used for X‐ray dose measurements during diagnostic examinations and have begun to be used for three‐dimensional X‐ray dose measurements using the high‐resolution characteristics of Gafchromic films for computed tomography. However, the problem of unevenness in Gafchromic film active layers needs to be resolved. Double exposures using X‐rays are performed during therapeutic radiology, although this is difficult for a diagnostic examination because of a heel effect. Thus, it has been suggested that ultraviolet (UV) radiation be used as a substitute for X‐rays. However, the appropriate UV wavelength has not been determined. Thus, we conducted this study to decide an appropriate UV wavelength. UV peak wavelengths of 245 nm (UV‐A), 310 nm (UV‐B), and 365 nm (UV‐C) were used to irradiate EBT2 and EBT3 films. Each UV wavelength was irradiated for 5, 15, 30, and 60 min, and irradiation was then repeated every 60 min up to 360 min. Gafchromic films were scanned after every irradiation using a flatbed scanner. Images were split into RGB images, and red images were analyzed using ImageJ, version 1.44, image analysis software. A region of interest (ROI) one‐half inch in diameter was placed in the center of subtracted Gafchromic film images, and UV irradiation times were plotted against mean pixel values. There were reactions in the front and back of Gafchromic EBT3 and the back of Gafchromic EBT2 with UV‐A and UV‐B. However, UV‐C resulted in some reactions in both sides of Gafchromic EBT2 and EBT3. The UV‐A and UV‐B wavelengths should be used. PACS number(s): 87.53 Bn PMID:26699578

  2. Correction of nonuniformity error of Gafchromic EBT2 and EBT3.

    PubMed

    Katsuda, Toshizo; Gotanda, Rumi; Gotanda, Tatsuhiro; Akagawa, Takuya; Tanki, Nobuyoshi; Kuwano, Tadao; Yabunaka, Kouichi

    2016-05-08

    This study investigates an X-ray dose measurement method for computed tomography using Gafchromic films. Nonuniformity of the active layer is a major problem in Gafchromic films. In radiotherapy, nonuniformity error is reduced by applying the double-exposure technique, but this is impractical in diagnostic radiology because of the heel effect. Therefore, we propose replacing the X-rays in the double-exposure technique with ultraviolet (UV)-A irradiation of Gafchromic EBT2 and EBT3. To improve the reproducibility of the scan position, Gafchromic EBT2 and EBT3 films were attached to a 3-mm-thick acrylic plate. The samples were then irradiated with a 10 W UV-A fluorescent lamp placed at a distance of 72cm for 30, 60, and 90 minutes. The profile curves were evaluated along the long and short axes of the film center, and the standard deviations of the pixel values were calculated over large areas of the films. Paired t-test was performed. UV-A irradiation exerted a significant effect on Gafchromic EBT2 (paired t-test; p = 0.0275) but not on EBT3 (paired t-test; p = 0.2785). Similarly, the homogeneity was improved in Gafchromic EBT2 but not in EBT3. Therefore, the double-exposure technique under UV-A irradiation is suitable only for EBT2 films.

  3. Correction of nonuniformity error of Gafchromic EBT2 and EBT3

    PubMed Central

    Gotanda, Rumi; Gotanda, Tatsuhiro; Akagawa, Takuya; Tanki, Nobuyoshi; Kuwano, Tadao; Yabunaka, Kouichi

    2016-01-01

    This study investigates an X‐ray dose measurement method for computed tomography using Gafchromic films. Nonuniformity of the active layer is a major problem in Gafchromic films. In radiotherapy, nonuniformity error is reduced by applying the double‐exposure technique, but this is impractical in diagnostic radiology because of the heel effect. Therefore, we propose replacing the X‐rays in the double‐exposure technique with ultraviolet (UV)‐A irradiation of Gafchromic EBT2 and EBT3. To improve the reproducibility of the scan position, Gafchromic EBT2 and EBT3 films were attached to a 3‐mm‐thick acrylic plate. The samples were then irradiated with a 10 W UV‐A fluorescent lamp placed at a distance of 72 cm for 30, 60, and 90 minutes. The profile curves were evaluated along the long and short axes of the film center, and the standard deviations of the pixel values were calculated over large areas of the films. Paired t‐test was performed. UV‐A irradiation exerted a significant effect on Gafchromic EBT2 (paired t‐test; p=0.0275) but not on EBT3 (paired t‐test; p=0.2785). Similarly, the homogeneity was improved in Gafchromic EBT2 but not in EBT3. Therefore, the double‐exposure technique under UV‐A irradiation is suitable only for EBT2 films. PACS number(s): 87.53 Bn PMID:27167258

  4. Evaluation of the magnitude of EBT Gafchromic film polarization effects.

    PubMed

    Butson, M J; Cheung, T; Yu, P K N

    2009-03-01

    Gafchromic EBT film, has become a main dosimetric tools for quantitative evaluation of radiation doses in radiation therapy application. One aspect of variability using EBT Gafchromic film is the magnitude of the orientation effect when analysing the film in landscape or portrait mode. This work has utilized a > 99% plane polarized light source and a non-polarized diffuse light source to investigate the absolute magnitude of EBT Gafchromic films polarization or orientation effects. Results have shown that using a non-polarized light source produces a negligible orientation effect for EBT Gafchromic film and thus the angle of orientation is not important. However, the film exhibits a significant variation in transmitted optical density with angle of orientation to polarized light producing more than 100% increase, or over a doubling of measured OD for films irradiated with x-rays up to dose levels of 5 Gy. The maximum optical density was found to be in a plane at an angle of 14 degrees +/- 7 degrees (2 SD) when the polarizing sheet is turned clockwise with respect to the film. As the magnitude of the orientation effect follows a sinusoidal shape it becomes more critical for alignment accuracy of the film with respect to the polarizing direction in the anticlockwise direction as this will place the alignment of the polarizing axes on the steeper gradient section of the sinusoidal pattern. An average change of 4.5% per 5 degrees is seen for an anticlockwise polarizer rotation where as the effect is 1.2% per 5 degrees for an clockwise polarizer rotation. This may have consequences to the positional accuracy of placement of the EBT Gafchromic film on a scanner as even a 1 degree alignment error can cause an approximate 1% error in analysis. The magnitude of the orientation effect is therefore dependant on the degree of polarization of the scanning light source and can range from negligible (diffuse LED light source) through to more than 100% or doubling of OD variation

  5. A study on comparison of Gafchromic EBT2 film response under single and cumulative exposure conditions

    PubMed Central

    Ganapathy, K.; Kurup, P.G.G.; Murali, V.; Muthukumaran, M.; Velmurugan, J.

    2013-01-01

    Gafchromic films are used as dosimeter for in vivo and in phantom dose measurements. The dose response of Gafchromic EBT2 film under single and repeated exposure conditions is compared in this study to analyze the usability of Gafchromic EBT2 films in cumulative dose measurements. The post-irradiation change in response of the film is studied for up to 4 days after irradiation. The effect of repeated exposure to scanner light on the response of the film is also studied. To check usability of Gafchromic EBT2 films in cumulative dose measurements, three EBT2 films were exposed to a daily fraction dose of 100 cGy, 150 cGy and 200 cGy, respectively, for 4 days. The dose response of the films exposed to cumulative irradiation was compared with the dose measured from films exposed to the same dose but in a single exposure. It is observed that the post-irradiation darkening of the film does not saturate and continue to take place even 4 days after irradiation. The dose measured from the EBT2 films after 4 days from irradiation was around 2% higher than the dose measured from the same films at 24 hours post-irradiation. It was also observed that the repeated exposure to scanner light does not produce any significant change in the film response. The dose response of films exposed to cumulative irradiation agrees with the dose response of films exposed to the same dose in a single irradiation with less than 3% difference. Gafchromic EBT2 films can be used to measure the cumulative dose delivered over multiple fractions, when the delivered dose is uniform across the film. PMID:24672151

  6. Technical Note: On GAFChromic EBT-XD film and the lateral response artifact.

    PubMed

    Lewis, David F; Chan, Maria F

    2016-02-01

    The new radiochromic film, GAFChromic EBT-XD, contains the same active material, lithium-10,12-pentacosadiynoate, as GAFChromic EBT3, but the crystalline form is different. This work investigates the effect of this change on the well-known lateral response artifact when EBT-XD film is digitized on a flatbed scanner. The dose response of a single production lot of EBT-XD was characterized by scanning an unexposed film plus a set of films exposed to doses between 2.5 and 50 Gy using 6 MV photons. To characterize the lateral response artifact, the authors used the unexposed film plus a subset of samples exposed to doses between 20 and 50 Gy. Digital images of these films were acquired at seven discrete lateral locations perpendicular to the scan direction on three Epson 10000XL scanners. Using measurements at the discrete lateral positions, the scanner responses were determined as a function of the lateral position of the film. From the data for each scanner, a set of coefficients were derived whereby measured response values could be corrected to remove the effects of the lateral response artifact. The EBT-XD data were analyzed as in their previous work and compared to results reported for EBT3 in that paper. For films scanned in the same orientation and having equal responses, the authors found that the lateral response artifact for EBT-XD and EBT3 films was remarkably similar. For both films, the artifact increases with increased net response. However, as EBT-XD is less sensitive than EBT3, a greater exposure dose is required to reach the same net response. On this basis, the lower sensitivity of EBT-XD relative to EBT3 results in less net response change for equal exposure and a reduction in the impact of the lateral response artifact. The shape of the crystalline active component in EBT-XD and EBT3 does not affect the fundamental existence of the lateral response artifact when the films are digitized on flatbed scanners. Owing its lower sensitivity, EBT-XD film

  7. Technical Note: On GAFChromic EBT-XD film and the lateral response artifact

    PubMed Central

    Lewis, David F.

    2016-01-01

    Purpose: The new radiochromic film, GAFChromic EBT-XD, contains the same active material, lithium-10,12-pentacosadiynoate, as GAFChromic EBT3, but the crystalline form is different. This work investigates the effect of this change on the well-known lateral response artifact when EBT-XD film is digitized on a flatbed scanner. Methods: The dose response of a single production lot of EBT-XD was characterized by scanning an unexposed film plus a set of films exposed to doses between 2.5 and 50 Gy using 6 MV photons. To characterize the lateral response artifact, the authors used the unexposed film plus a subset of samples exposed to doses between 20 and 50 Gy. Digital images of these films were acquired at seven discrete lateral locations perpendicular to the scan direction on three Epson 10000XL scanners. Using measurements at the discrete lateral positions, the scanner responses were determined as a function of the lateral position of the film. From the data for each scanner, a set of coefficients were derived whereby measured response values could be corrected to remove the effects of the lateral response artifact. The EBT-XD data were analyzed as in their previous work and compared to results reported for EBT3 in that paper. Results: For films scanned in the same orientation and having equal responses, the authors found that the lateral response artifact for EBT-XD and EBT3 films was remarkably similar. For both films, the artifact increases with increased net response. However, as EBT-XD is less sensitive than EBT3, a greater exposure dose is required to reach the same net response. On this basis, the lower sensitivity of EBT-XD relative to EBT3 results in less net response change for equal exposure and a reduction in the impact of the lateral response artifact. Conclusions: The shape of the crystalline active component in EBT-XD and EBT3 does not affect the fundamental existence of the lateral response artifact when the films are digitized on flatbed scanners

  8. Determining optimum wavelength of ultraviolet rays to pre-exposure of non-uniformity error correction in Gafchromic EBT2 films

    NASA Astrophysics Data System (ADS)

    Katsuda, Toshizo; Gotanda, Rumi; Gotanda, Tatsuhiro; Akagawa, Takuya; Tanki, Nobuyoshi; Kuwano, Tadao; Noguchi, Atsushi; Yabunaka, Kouichi

    2018-03-01

    Gafchromic films have been used to measure X-ray doses in diagnostic radiology such as computed tomography. The double-exposure technique is used to correct non-uniformity error of Gafchromic EBT2 films. Because of the heel effect of diagnostic x-rays, ultraviolet A (UV-A) is intended to be used as a substitute for x-rays. When using a UV-A light-emitting diode (LED), it is necessary to determine the effective optimal UV wavelength for the active layer of Gafchromic EBT2 films. This study evaluated the relation between the increase in color density of Gafchromic EBT2 films and the UV wavelengths. First, to correct non-uniformity, a Gafchromic EBT2 film was pre-irradiated using uniform UV-A radiation for 60 min from a 72-cm distance. Second, the film was irradiated using a UV-LED with a wavelength of 353-410 nm for 60 min from a 5.3-cm distance. The maximum, minimum, and mean ± standard deviation (SD) of pixel values of the subtraction images were evaluated using 0.5 inches of a circular region of interest (ROI). The highest mean ± SD (8915.25 ± 608.86) of the pixel value was obtained at a wavelength of 375 nm. The results indicated that 375 nm is the most effective and sensitive wavelength of UV-A for Gafchromic EBT2 films and that UV-A can be used as a substitute for x-rays in the double-exposure technique.

  9. Measuring diffuse ultraviolet exposures using Gafchromic EBT3 films

    NASA Astrophysics Data System (ADS)

    Tsang, P. Y.; Chan, P. M.; Yu, K. N.

    The present work proposed to use Gafchromic EBT3 films for determining spatial variations of ultraviolet (UV) exposures in the environment, which required many simultaneous measurements, and demonstrated the feasibility by showing the variation of the diffuse component of solar (UVA + UVB) exposures (in Jcm-2) in shaded areas under overhangs with the elevation angle θ for the edge of the overhangs.

  10. Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry.

    PubMed

    Desroches, Joannie; Bouchard, Hugo; Lacroix, Frédéric

    2010-04-01

    The purpose of this study is to determine the effect on the measured optical density of scanning on either side of a Gafchromic EBT and EBT2 film using an Epson (Epson Canada Ltd., Toronto, Ontario) 10000XL flat bed scanner. Calibration curves were constructed using EBT2 film scanned in landscape orientation in both reflection and transmission mode on an Epson 10000XL scanner. Calibration curves were also constructed using EBT film. Potential errors due to an optical density difference from scanning the film on either side ("face up" or "face down") were simulated. Scanning the film face up or face down on the scanner bed while keeping the film angular orientation constant affects the measured optical density when scanning in reflection mode. In contrast, no statistically significant effect was seen when scanning in transmission mode. This effect can significantly affect relative and absolute dose measurements. As an application example, the authors demonstrate potential errors of 17.8% by inverting the film scanning side on the gamma index for 3%-3 mm criteria on a head and neck intensity modulated radiotherapy plan, and errors in absolute dose measurements ranging from 10% to 35% between 2 and 5 Gy. Process consistency is the key to obtaining accurate and precise results in Gafchromic film dosimetry. When scanning in reflection mode, care must be taken to place the film consistently on the same side on the scanner bed.

  11. Quantitative investigation of the effects of the scanning parameters in the digitization of EBT and EBT2 Gafchromic film dosimetry with flatbed scanners.

    PubMed

    Hu, Yida; Ahmad, Salahuddin; Ali, Imad

    2012-01-01

    With increasing popularity and complexity of intensity-modulated radiation therapy (IMRT) delivery modalities including regular and arc therapies, there is a growing challenge for validating the accuracy of dose distributions. Gafchromic films have superior characteristics for dose verification over other conventional dosimeters. In order to optimize the use of Gafchromic films in clinical IMRT quality assurance procedures, the scanning parameters of EBT and EBT2 films with a flatbed scanner were investigated. The effects of several parameters including scanning position, orientation, uniformity, film sensitivity and optical density (OD) growth after irradiation were quantified. The profiles of the EBT and EBT2 films had a noise level of 0.6% and 0.7%, respectively. Considerable orientation dependence was observed and the scanner value difference between landscape and portrait modes were about 12% and 10% for EBT and EBT2 films, respectively. The highest response sensitivity was observed using digitized red color images of the EBT2 film scanned with landscape mode. The total system non-uniformity composed of contributions from the film and the scanner was less than 1.7%. OD variations showed that EBT gray scale grew slower, however, reached higher growth values of 15% when compared with EBT2 gray scale which grew 12% after a long time (480 hours) post-irradiation. The EBT film using the red color channel showed the minimal growth where OD increased up to 3% within 3 days after irradiation, and took one week to stabilize.

  12. Evaluation of Gafchromic EBT-XD film, with comparison to EBT3 film, and application in high dose radiotherapy verification.

    PubMed

    Palmer, Antony L; Dimitriadis, Alexis; Nisbet, Andrew; Clark, Catharine H

    2015-11-21

    There is renewed interest in film dosimetry for the verification of dose delivery of complex treatments, particularly small fields, compared to treatment planning system calculations. A new radiochromic film, Gafchromic EBT-XD, is available for high-dose treatment verification and we present the first published evaluation of its use. We evaluate the new film for MV photon dosimetry, including calibration curves, performance with single- and triple-channel dosimetry, and comparison to existing EBT3 film. In the verification of a typical 25 Gy stereotactic radiotherapy (SRS) treatment, compared to TPS planned dose distribution, excellent agreement was seen with EBT-XD using triple-channel dosimetry, in isodose overlay, maximum 1.0 mm difference over 200-2400 cGy, and gamma evaluation, mean passing rate 97% at 3% locally-normalised, 1.5 mm criteria. In comparison to EBT3, EBT-XD gave improved evaluation results for the SRS-plan, had improved calibration curve gradients at high doses, and had reduced lateral scanner effect. The dimensions of the two films are identical. The optical density of EBT-XD is lower than EBT3 for the same dose. The effective atomic number for both may be considered water-equivalent in MV radiotherapy. We have validated the use of EBT-XD for high-dose, small-field radiotherapy, for routine QC and a forthcoming multi-centre SRS dosimetry intercomparison.

  13. Evaluation of Gafchromic EBT-XD film, with comparison to EBT3 film, and application in high dose radiotherapy verification

    NASA Astrophysics Data System (ADS)

    Palmer, Antony L.; Dimitriadis, Alexis; Nisbet, Andrew; Clark, Catharine H.

    2015-11-01

    There is renewed interest in film dosimetry for the verification of dose delivery of complex treatments, particularly small fields, compared to treatment planning system calculations. A new radiochromic film, Gafchromic EBT-XD, is available for high-dose treatment verification and we present the first published evaluation of its use. We evaluate the new film for MV photon dosimetry, including calibration curves, performance with single- and triple-channel dosimetry, and comparison to existing EBT3 film. In the verification of a typical 25 Gy stereotactic radiotherapy (SRS) treatment, compared to TPS planned dose distribution, excellent agreement was seen with EBT-XD using triple-channel dosimetry, in isodose overlay, maximum 1.0 mm difference over 200-2400 cGy, and gamma evaluation, mean passing rate 97% at 3% locally-normalised, 1.5 mm criteria. In comparison to EBT3, EBT-XD gave improved evaluation results for the SRS-plan, had improved calibration curve gradients at high doses, and had reduced lateral scanner effect. The dimensions of the two films are identical. The optical density of EBT-XD is lower than EBT3 for the same dose. The effective atomic number for both may be considered water-equivalent in MV radiotherapy. We have validated the use of EBT-XD for high-dose, small-field radiotherapy, for routine QC and a forthcoming multi-centre SRS dosimetry intercomparison.

  14. Technical Note: Initial characterization of the new EBT-XD Gafchromic film.

    PubMed

    Grams, Michael P; Gustafson, Jon M; Long, Kenneth M; de los Santos, Luis E Fong

    2015-10-01

    To assess the dosimetric accuracy and energy dependence of the new EBT-eXtended Dose (XD) Gafchromic film and to compare the lateral response artifact (LRA) between EBT-XD and EBT3 film. EBT3 and EBT-XD calibration curves were created by exposing films to known doses from 0 to 3000 cGy using a 6 MV beam. To assess the accuracy and dynamic range of EBT-XD, a 60° enhanced dynamic wedge (EDW) was used to deliver a dose range of approximately 200-2900 cGy. Comparison to treatment planning system (TPS) calculation was made using a gamma analysis with 2%/2 mm passing criteria. To assess and compare the LRA between EBT3 and EBT-XD, 21 × 21 cm(2) open fields delivered doses of 1000, 2000, and 3000 cGy to both types of film. Films were placed at the center of the scanner, and ratios of measured to TPS predicted doses were calculated at 50 and 80 mm lateral from the scanner center in order to quantitatively assess the LRA. To evaluate the energy dependence of EBT-XD film, seven known doses ranging from 400 to 3000 cGy were delivered using both 6 and 18 MV beams and the resulting optical densities (ODs) compared. The gamma passing rate was 99.1% for the 6 MV EDW delivery. EBT-XD film exhibited minimal LRA (<1%) up to 3000 cGy. In contrast, EBT3 demonstrated an under-response of 11.3% and 22.7% at lateral positions of 50 and 80 mm, respectively, for the 3000 cGy exposure. Differences between ODs of the EBT-XD films exposed to known doses from 6 to 18 MV beams were <0.8% suggesting minimal energy dependence throughout this energy range. The LRA of EBT-XD is greatly reduced when compared to EBT3. This in combination with its accuracy from 0 to 3000 cGy and minimal energy dependence from 6 to 18 MV makes EBT-XD film well suited for dosimetric measurements in high dose SRS/SBRT applications.

  15. Dose distribution verification for GYN brachytherapy using EBT Gafchromic film and TG-43 calculation.

    PubMed

    Gholami, Somayeh; Mirzaei, Hamid Reza; Jabbary Arfaee, Ali; Jaberi, Ramin; Nedaie, Hassan Ali; Rabi Mahdavi, Seied; Rajab Bolookat, Eftekhar; Meigooni, Ali S

    2016-01-01

    Verification of dose distributions for gynecological (GYN) brachytherapy implants using EBT Gafchromic film. One major challenge in brachytherapy is to verify the accuracy of dose distributions calculated by a treatment planning system. A new phantom was designed and fabricated using 90 slabs of 18 cm × 16 cm × 0.2 cm Perspex to accommodate a tandem and Ovoid assembly, which is normally used for GYN brachytherapy treatment. This phantom design allows the use of EBT Gafchromic films for dosimetric verification of GYN implants with a cobalt-60 HDR system or a LDR Cs-137 system. Gafchromic films were exposed using a plan that was designed to deliver 1.5 Gy of dose to 0.5 cm distance from the lateral surface of ovoids from a pair of ovoid assembly that was used for treatment vaginal cuff. For a quantitative analysis of the results for both LDR and HDR systems, the measured dose values at several points of interests were compared with the calculated data from a commercially available treatment planning system. This planning system was utilizing the TG-43 formalism and parameters for calculation of dose distributions around a brachytherapy implant. The results of these investigations indicated that the differences between the calculated and measured data at different points were ranging from 2.4% to 3.8% for the LDR Cs-137 and HDR Co-60 systems, respectively. The EBT Gafchromic films combined with the newly designed phantom could be utilized for verification of the dose distributions around different GYN implants treated with either LDR or HDR brachytherapy procedures.

  16. Quality assurance of alpha-particle dosimetry using peeled-off Gafchromic EBTfilm

    NASA Astrophysics Data System (ADS)

    Ng, C. Y. P.; Chun, S. L.; Yu, K. N.

    2016-08-01

    A novel alpha-particle dosimetry technique using Gafchromic EBT3 film has recently been proposed for calibrating the activity of alpha-emitting radiopharmaceuticals. In the present paper, we outlined four measures which could further help assure the quality of the method. First, we suggested an alternative method in fabricating the peeled-off EBT3 film. Films with a chosen size were cut from the original films and all the edges were sealed with silicone. These were immersed into deionized water for 19 d and the polyester covers of the EBT3 films could then be easily peeled off. The active layers in these peeled-off EBT3 films remained intact, and these films could be prepared reproducibly with ease. Second, we proposed a check on the integrity of the peeled-off film by comparing the responses of the pristine and peeled-off EBT3 films to the same X-ray irradiation. Third, we highlighted the importance of scanning directions of the films. The ;landscape; and ;portrait; scanning directions were defined as the scanning directions perpendicular and parallel to the long edge of the original EBT3 films, respectively. Our results showed that the responses were different for different scanning directions. As such, the same scanning direction should be used every time. Finally, we cautioned the need to confirm the uniformity of the alpha-particle source used for calibration. Radiochromic films are well known for their capability of providing two-dimensional dosimetric information. As such, EBT3 films could also be conveniently used to check the uniformity of the alpha-particle source.

  17. Calibrating page sized Gafchromic EBT3 films

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Crijns, W.; Maes, F.; Heide, U. A. van der

    2013-01-15

    Purpose: The purpose is the development of a novel calibration method for dosimetry with Gafchromic EBT3 films. The method should be applicable for pretreatment verification of volumetric modulated arc, and intensity modulated radiotherapy. Because the exposed area on film can be large for such treatments, lateral scan errors must be taken into account. The correction for the lateral scan effect is obtained from the calibration data itself. Methods: In this work, the film measurements were modeled using their relative scan values (Transmittance, T). Inside the transmittance domain a linear combination and a parabolic lateral scan correction described the observed transmittancemore » values. The linear combination model, combined a monomer transmittance state (T{sub 0}) and a polymer transmittance state (T{sub {infinity}}) of the film. The dose domain was associated with the observed effects in the transmittance domain through a rational calibration function. On the calibration film only simple static fields were applied and page sized films were used for calibration and measurements (treatment verification). Four different calibration setups were considered and compared with respect to dose estimation accuracy. The first (I) used a calibration table from 32 regions of interest (ROIs) spread on 4 calibration films, the second (II) used 16 ROIs spread on 2 calibration films, the third (III), and fourth (IV) used 8 ROIs spread on a single calibration film. The calibration tables of the setups I, II, and IV contained eight dose levels delivered to different positions on the films, while for setup III only four dose levels were applied. Validation was performed by irradiating film strips with known doses at two different time points over the course of a week. Accuracy of the dose response and the lateral effect correction was estimated using the dose difference and the root mean squared error (RMSE), respectively. Results: A calibration based on two films was the

  18. UNLAMINATED GAFCHROMIC EBT3 FILM FOR ULTRAVIOLET RADIATION MONITORING.

    PubMed

    Welch, David; Randers-Pehrson, Gerhard; Spotnitz, Henry M; Brenner, David J

    2017-11-01

    Measurement of ultraviolet (UV) radiation is important for human health, especially with the expanded usage of short wavelength UV for sterilization purposes. This work examines unlaminated Gafchromic EBT3 film for UV radiation monitoring. The authors exposed the film to select wavelengths in the UV spectrum, ranging from 207 to 328 nm, and measured the change in optical density. The response of the film is wavelength dependent, and of the wavelengths tested, the film was most sensitive to 254 nm light, with measurable values as low as 10 µJ/cm2. The film shows a dose-dependent response that extends over more than four orders of magnitude. The response of the film to short wavelength UV is comparable to the daily safe exposure limits for humans, thus making it valuable as a tool for passive UV radiation monitoring. © The Author 2017. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  19. Evaluation of the Gafchromic{sup Registered-Sign} EBT2 film for the dosimetry of radiosurgical beams

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Larraga-Gutierrez, Jose M.; Garcia-Hernandez, Diana; Garcia-Garduno, Olivia A.

    2012-10-15

    Purpose: Radiosurgery uses small fields and high-radiation doses to treat intra- and extracranial lesions in a single session. The lack of a lateral electronic equilibrium and the presence of high-dose gradients in these fields are challenges for adequate measurements. The availability of radiation detectors with the high spatial resolution required is restricted to only a few. Stereotactic diodes and EBT radiochromic films have been demonstrated to be good detectors for small-beam dosimetry. Because the stereotactic diode is the standard measurement for the dosimetry of radiosurgical beams, the goal of this work was to perform measurements with the radiochromic film Gafchromic{supmore » Registered-Sign} EBT2 and compare its results with a stereotactic diode. Methods: Total scatter factors, tissue maximum, and off-axis ratios from a 6 MV small photon beams were measured using EBT2 radiochromic film in a water phantom. The film-measured data were evaluated by comparing it with the data measured with a stereotactic field diode (IBA-Dosimetry). Results: The film and diode measurements had excellent agreement. The differences between the detectors were less than or equal to 2.0% for the tissue maximum and the off-axis ratios. However, for the total scatter factors, there were significant differences, up to 4.9% (relative to the reference field), for field sizes less than 1.0 cm. Conclusions: This work found that the Gafchromic{sup Registered-Sign} EBT2 film is adequate for small photon beam measurements, particularly for tissue maximum and off-axis ratios. However, careful attention must be taken when measuring output factors of small beams below 1.0 cm due to the film's energy dependence. The measurement differences may be attributable to the film's active layer composition because EBT2 incorporates higher Z elements (i.e., bromide and potassium), hence revealing a potential energy dependence for the dosimetry of small photon beams.« less

  20. SU-E-T-128: Dosimetric Characteristics of Gafchromic EBT3 Films for Megavoltage Photon and Proton Beams.

    PubMed

    Hayashi, N; Kato, H; Yada, R; Matsunaga, T; Ehara, I; Adachi, Y

    2012-06-01

    Gafchromic film for quantitative analysis was renewed from EBT2 to EBT3 film in November 2011. The purpose of this study is to investigate the relevant characteristics of EBT3 film for its application in dosimetric verification for IMRT/VMAT or proton therapy. We investigated the characteristics of EBT3 film with comparison of previous EBT2 film. The experiments in this study composed two categories. At first, the photo spectroscopy for the irradiated film was compared between EBT2 and EBT3. The film 1 day after the irradiation was analyzed by a photo spectrometer (SR520: JASCO Corporation, Japan). Secondly, we investigated several calibration curves which obtained by same batch. The films were calibrated by irradiation the films to 13 dose steps. The irradiated films were scanned by a flatbed scanner (ES-10000XL, Epson-Seiko Corporation, Japan). The difference on scan orientation was evaluated alternate portrait and landscape directions. The photon and proton beams were delivered from Clinac 21EX (Varian) and Mitsubishi machine, respectively. The peak absorption wavelength of EBT3 film and its response at all active range were basically same with that of EBT2 film. The peak wavelength of photo absorption in EBT3 was observed at 585 and 634 nm. The fog optical density was increased due to the hazy matte polyester for active layer. However, there is no change the tendency of the calibration curve responding to megavoltage photon and proton beams. The scan orientation dependency of EBT3 film was observed with similar to EBT2 film. The optical density of portrait orientation was 10% higher than that of landscape orientation. The dosimetric characteristics of EBT3 film were basically same with EBT2 film. With regard to the matte polyester, the creation of Newton's rings during scanning procedure was reduced. However, the suitable scan protocol should be used for accurate film dosimetry. © 2012 American Association of Physicists in Medicine.

  1. A study of surface dosimetry for breast cancer radiotherapy treatments using Gafchromic EBT2 film

    PubMed Central

    Hill, Robin F.; Whitaker, May; Kim, Jung‐Ha; Kuncic, Zdenka

    2012-01-01

    The present study quantified surface doses on several rectangular phantom setups and on curved surface phantoms for a 6 MV photon field using the Attix parallel‐plate chamber and Gafchromic EBT2 film. For the rectangular phantom setups, the surface doses on a homogenous water equivalent phantom and a water equivalent phantom with 60 mm thick lung equivalent material were measured. The measurement on the homogenous phantom setup showed consistency in surface and near‐surface doses between an open field and enhanced dynamic wedge (EDW) fields, whereas physical wedged fields showed small differences. Surface dose measurements made using the EBT2 film showed good agreement with results of the Attix chamber and results obtained in previous studies which used other dosimeters within the measurement uncertainty of 3.3%. The surface dose measurements on the phantom setup with lung equivalent material showed a small increase without bolus and up to 6.9% increase with bolus simulating the increase of chest wall thickness. Surface doses on the cylindrical CT phantom and customized Perspex chest phantom were measured using the EBT2 film with and without bolus. The results indicate the important role of the presence of bolus if the clinical target volume (CTV) is quite close to the surface. Measurements on the cylindrical phantom suggest that surface doses at the oblique positions of 60° and 90° are mainly caused by the lateral scatter from the material inside the phantom. In the case of a single tangential irradiation onto Perspex chest phantom, the distribution of the surface dose with and without bolus materials showed opposing inclination patterns, whereas the dose distribution for two opposed tangential fields gave symmetric dose distribution. This study also demonstrates the suitability of Gafchromic EBT2 film for surface dose measurements in megavoltage photon beams. PACS number: 87.53.Bn PMID:22584169

  2. Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification.

    PubMed

    Sankar, A; Ayyangar, Komanduri M; Nehru, R Mothilal; Kurup, P G Gopalakrishna; Murali, V; Enke, Charles A; Velmurugan, J

    2006-01-01

    The quantitative dose validation of intensity-modulated radiation therapy (IMRT) plans require 2-dimensional (2D) high-resolution dosimetry systems with uniform response over its sensitive region. The present work deals with clinical use of commercially available self-developing Radio Chromic Film, Gafchromic EBT film, for IMRT dose verification. Dose response curves were generated for the films using a VXR-16 film scanner. The results obtained with EBT films were compared with the results of Kodak extended dose range 2 (EDR2) films. The EBT film had a linear response between the dose range of 0 to 600 cGy. The dose-related characteristics of the EBT film, such as post irradiation color growth with time, film uniformity, and effect of scanning orientation, were studied. There was up to 8.6% increase in the color density between 2 to 40 hours after irradiation. There was a considerable variation, up to 8.5%, in the film uniformity over its sensitive region. The quantitative differences between calculated and measured dose distributions were analyzed using DTA and Gamma index with the tolerance of 3% dose difference and 3-mm distance agreement. The EDR2 films showed consistent results with the calculated dose distributions, whereas the results obtained using EBT were inconsistent. The variation in the film uniformity limits the use of EBT film for conventional large-field IMRT verification. For IMRT of smaller field sizes (4.5 x 4.5 cm), the results obtained with EBT were comparable with results of EDR2 films.

  3. Comparison of Gafchromic EBT2 and EBT3 for patient-specific quality assurance: Cranial stereotactic radiosurgery using volumetric modulated arc therapy with multiple noncoplanar arcs

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fiandra, Christian; Fusella, Marco; Filippi, Andrea Riccardo

    2013-08-15

    Purpose: Patient-specific quality assurance in volumetric modulated arc therapy (VMAT) brain stereotactic radiosurgery raises specific issues on dosimetric procedures, mainly represented by the small radiation fields associated with the lack of lateral electronic equilibrium, the need of small detectors and the high dose delivered (up to 30 Gy). Gafchromic{sup TM} EBT2 and EBT3 films may be considered the dosimeter of choice, and the authors here provide some additional data about uniformity correction for this new generation of radiochromic films.Methods: A new analysis method using blue channel for marker dye correction was proposed for uniformity correction both for EBT2 and EBT3more » films. Symmetry, flatness, and field-width of a reference field were analyzed to provide an evaluation in a high-spatial resolution of the film uniformity for EBT3. Absolute doses were compared with thermoluminescent dosimeters (TLD) as baseline. VMAT plans with multiple noncoplanar arcs were generated with a treatment planning system on a selected pool of eleven patients with cranial lesions and then recalculated on a water-equivalent plastic phantom by Monte Carlo algorithm for patient-specific QA. 2D quantitative dose comparison parameters were calculated, for the computed and measured dose distributions, and tested for statistically significant differences.Results: Sensitometric curves showed a different behavior above dose of 5 Gy for EBT2 and EBT3 films; with the use of inhouse marker-dye correction method, the authors obtained values of 2.5% for flatness, 1.5% of symmetry, and a field width of 4.8 cm for a 5 × 5 cm{sup 2} reference field. Compared with TLD and selecting a 5% dose tolerance, the percentage of points with ICRU index below 1 was 100% for EBT2 and 83% for EBT3. Patients analysis revealed statistically significant differences (p < 0.05) between EBT2 and EBT3 in the percentage of points with gamma values <1 (p= 0.009 and p= 0.016); the percent difference as

  4. EXPERIMENTAL EVALUATION OF DOSIMETRIC CHARACTERIZATION OF GAFCHROMIC EBT3 AND EBT-XD FILMS FOR CLINICAL CARBON ION BEAMS.

    PubMed

    Yonai, Shunsuke; Arai, Chinatsu; Shimoyama, Kaoru; Fournier-Bidoz, Nathalie

    2018-02-03

    Radiochromic film is a very useful tool for 2D dosimetric measurements in radiotherapy because it is self-developing and has very high-spatial resolution. However, considerable care has to be taken in ion beam radiotherapy owing to the quenching effect of high-linear energy transfer (LET) radiation. In this study, the dose responses of GAFchromic EBT3 and EBT-XD films were experimentally investigated using the clinical carbon ion beam at the Heavy Ion Medical Accelerator in Chiba. Results showed that the relations between absorbed dose and net optical density could be expressed well using an equation proposed by Reinhardt (2015). The quenching effect was evaluated by determining their relative efficiencies for photon irradiation as a function of LET. A correction equation derived in this study allowed the absorbed dose to be determined in the small irradiation field used for carbon ion radiotherapy eye treatments. This study contributes to establishing an absolute dosimetry procedure for heavy ion beams using radiochromic film. © The Author(s) 2018. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  5. SU-E-T-96: Energy Dependence of the New GafChromic- EBT3 Film's Dose Response-Curve.

    PubMed

    Chiu-Tsao, S; Massillon-Jl, G; Domingo-Muñoz, I; Chan, M

    2012-06-01

    To study and compare the dose response curves of the new GafChromic EBT3 film for megavoltage and kilovoltage x-ray beams, with different spatial resolution. Two sets of EBT3 films (lot#A101711-02) were exposed to each x-ray beam (6MV, 15MV and 50kV) at 8 dose values (50-3200cGy). The megavoltage beams were calibrated per AAPM TG-51 protocol while the kilovoltage beam was calibrated following the TG-61 using an ionization chamber calibrated at NIST. Each film piece was scanned three consecutive times in the center of Epson 10000XL flatbed scanner in transmission mode, landscape orientation, 48-bit color at two separate spatial resolutions of 75 and 300 dpi. The data were analyzed using ImageJ and, for each scanned image, a region of interest (ROI) of 2×2cm 2 at the field center was selected to obtain the mean pixel value with its standard deviation in the ROI. For each energy, dose value and spatial resolution, the average netOD and its associated uncertainty were determined. The Student's t-test was performed to evaluate the statistical differences between the netOD/dose values of the three energy modalities, with different color channels and spatial resolutions. The dose response curves for the three energy modalities were compared in three color channels with 75 and 300dpi. Weak energy dependence was found. For doses above 100cGy, no statistical differences were observed between 6 and 15MV beams, regardless of spatial resolution. However, statistical differences were observed between 50kV and the megavoltage beams. The degree of energy dependence (from MV to 50kV) was found to be function of color channel, dose level and spatial resolution. The dose response curves for GafChromic EBT3 films were found to be weakly dependent on the energy of the photon beams from 6MV to 50kV. The degree of energy dependence varies with color channel, dose and spatial resolution. GafChromic EBT3 films were supplied by Ashland Corp. This work was partially supported by DGAPA

  6. High dose-per-pulse electron beam dosimetry: Usability and dose-rate independence of EBT3 Gafchromic films.

    PubMed

    Jaccard, Maud; Petersson, Kristoffer; Buchillier, Thierry; Germond, Jean-François; Durán, Maria Teresa; Vozenin, Marie-Catherine; Bourhis, Jean; Bochud, François O; Bailat, Claude

    2017-02-01

    The aim of this study was to assess the suitability of Gafchromic EBT3 films for reference dose measurements in the beam of a prototype high dose-per-pulse linear accelerator (linac), capable of delivering electron beams with a mean dose-rate (Ḋ m ) ranging from 0.07 to 3000 Gy/s and a dose-rate in pulse (Ḋ p ) of up to 8 × 10 6 Gy/s. To do this, we evaluated the overall uncertainties in EBT3 film dosimetry as well as the energy and dose-rate dependence of their response. Our dosimetric system was composed of EBT3 Gafchromic films in combination with a flatbed scanner and was calibrated against an ionization chamber traceable to primary standard. All sources of uncertainties in EBT3 dosimetry were carefully analyzed using irradiations at a clinical radiotherapy linac. Energy dependence was investigated with the same machine by acquiring and comparing calibration curves for three different beam energies (4, 8 and 12 MeV), for doses between 0.25 and 30 Gy. Ḋ m dependence was studied at the clinical linac by changing the pulse repetition frequency (f) of the beam in order to vary Ḋ m between 0.55 and 4.40 Gy/min, while Ḋ p dependence was probed at the prototype machine for Ḋ p ranging from 7 × 10 3 to 8 × 10 6 Gy/s. Ḋ p dependence was first determined by studying the correlation between the dose measured by films and the charge of electrons measured at the exit of the machine by an induction torus. Furthermore, we compared doses from the films to independently calibrated thermo-luminescent dosimeters (TLD) that have been reported as being dose-rate independent up to such high dose-rates. We report that uncertainty below 4% (k = 2) can be achieved in the dose range between 3 and 17 Gy. Results also demonstrated that EBT3 films did not display any detectable energy dependence for electron beam energies between 4 and 12 MeV. No Ḋ m dependence was found either. In addition, we obtained excellent consistency between films and TLDs over the entire Ḋ p

  7. Intensities of Incident and Transmitted Ultraviolet-A Rays through Gafchromic Films

    PubMed Central

    Katsuda, Toshizo; Gotanda, Rumi; Gotanda, Tatsuhiro; Akagawa, Takuya; Tanki, Nobuyoshi; Kuwano, Tadao; Noguchi, Atsushi; Yabunaka, Kouichi

    2017-01-01

    Gafchromic films have been applied to X-ray dosimetry in diagnostic radiology. To correct nonuniformity errors in Gafchromic films, X-rays in the double-exposure technique can be replaced with ultraviolet (UV)-A rays. Intensities of the incident and transmitted UV-A rays were measured. However, it is unclear whether the chemical color change of Gafchromic films affects the UV-A transmission intensity. Gafchromic EBT3 films were suitable to be used in this study because non-UV protection layers are present on both sides of the film. The film is placed between UV-A ray light-emitting diodes and a probe of a UV meter. Gafchromic EBT3 films were irradiated by UV-A rays for up to 60 min. Data for analysis were obtained in the subsequent 60 min. Images from before and after UV-A irradiation were subtracted. When using 375 nm UV-A, the mean ± standard deviation (SD) of the pixel values in the subtracted image was remarkably high (11,194.15 ± 586.63). However, the UV-A transmissivity remained constant throughout the 60 min irradiation period. The mean ± SD UV-A transmission intensity was 184.48 ± 0.50 μm/cm2. Our findings demonstrate that color density changes in Gafchromic EBT3 films do not affect their UV-A transmission. Therefore, Gafchromic films were irradiated by UV-A rays as a preexposure. PMID:28706354

  8. Gafchromic EBT-XD film: Dosimetry characterization in high-dose, volumetric-modulated arc therapy.

    PubMed

    Miura, Hideharu; Ozawa, Shuichi; Hosono, Fumika; Sumida, Naoki; Okazue, Toshiya; Yamada, Kiyoshi; Nagata, Yasushi

    2016-11-08

    Radiochromic films are important tools for assessing complex dose distributions. Gafchromic EBT-XD films have been designed for optimal performance in the 40-4,000 cGy dose range. We investigated the dosimetric characteristics of these films, including their dose-response, postexposure density growth, and dependence on scanner orientation, beam energy, and dose rate with applications to high-dose volumetric-modulated arc therapy (VMAT) verification. A 10 MV beam from a TrueBeam STx linear accelerator was used to irradiate the films with doses in the 0-4,000 cGy range. Postexposure coloration was analyzed at postirradiation times ranging from several minutes to 48 h. The films were also irradiated with 6 MV (dose rate (DR): 600 MU/min), 6 MV flattening filter-free (FFF) (DR: 1,400 MU/ min), and 10 MV FFF (DR: 2,400 MU/min) beams to determine the energy and dose-rate dependence. For clinical examinations, we compared the dose distribu-tion measured with EBT-XD films and calculated by the planning system for four VMAT cases. The red channel of the EBT-XD film exhibited a wider dynamic range than the green and blue channels. Scanner orientation yielded a variation of ~ 3% in the net optical density (OD). The difference between the film front and back scan orientations was negligible, with variation of ~ 1.3% in the net OD. The net OD increased sharply within the first 6 hrs after irradiation and gradually afterwards. No significant difference was observed for the beam energy and dose rate, with a variation of ~ 1.5% in the net OD. The gamma passing rates (at 3%, 3 mm) between the film- measured and treatment planning system (TPS)-calculated dose distributions under a high dose VMAT plan in the absolute dose mode were more than 98.9%. © 2016 The Authors.

  9. Gafchromic EBT3 film dosimetry in electron beams — energy dependence and improved film read‐out

    PubMed Central

    Ojala, Jarkko; Kaijaluoto, Sampsa; Jokelainen, Ilkka; Kosunen, Antti

    2016-01-01

    For megavoltage photon radiation, the fundamental dosimetry characteristics of Gafchromic EBT3 film were determined in  60Co gamma ray beam with addition of experimental and Monte Carlo (MC)‐simulated energy dependence of the film for 6 MV photon beam and 6 MeV, 9 MeV, 12 MeV, and 16 MeV electron beams in water phantom. For the film read‐out, two phase correction of scanner sensitivity was applied: a matrix correction for scanning area and dose‐dependent correction by iterative procedure. With these corrections, the uniformity of response can be improved to be within ±50 pixel values (PVs). To improve the read‐out accuracy, a procedure with flipped film orientations was established. With the method, scanner uniformity can be improved further and dust particles, scratches and/or dirt on scanner glass can be detected and eliminated. Responses from red and green channels were averaged for read‐out, which decreased the effect of noise present in values from separate channels. Since the signal level with the blue channel is considerably lower than with other channels, the signal variation due to different perturbation effects increases the noise level so that the blue channel is not recommended to be used for dose determination. However, the blue channel can be used for the detection of emulsion thickness variations for film quality evaluations with unexposed films. With electron beams ranging from 6 MeV to 16 MeV and at reference measurement conditions in water, the energy dependence of the EBT3 film is uniform within 0.5%, with uncertainties close to 1.6% (k=2). Including 6 MV photon beam and the electron beams mentioned, the energy dependence is within 1.1%. No notable differences were found between the experimental and MC‐simulated responses, indicating negligible change in intrinsic energy dependence of the EBT3 film for 6 MV photon beam and 6 MeV–16 MeV electron beams. Based on the dosimetric characteristics of the EBT3 film, the read

  10. The combination of the error correction methods of GAFCHROMIC EBT3 film

    PubMed Central

    Li, Yinghui; Chen, Lixin; Zhu, Jinhan; Liu, Xiaowei

    2017-01-01

    Purpose The aim of this study was to combine a set of methods for use of radiochromic film dosimetry, including calibration, correction for lateral effects and a proposed triple-channel analysis. These methods can be applied to GAFCHROMIC EBT3 film dosimetry for radiation field analysis and verification of IMRT plans. Methods A single-film exposure was used to achieve dose calibration, and the accuracy was verified based on comparisons with the square-field calibration method. Before performing the dose analysis, the lateral effects on pixel values were corrected. The position dependence of the lateral effect was fitted by a parabolic function, and the curvature factors of different dose levels were obtained using a quadratic formula. After lateral effect correction, a triple-channel analysis was used to reduce disturbances and convert scanned images from films into dose maps. The dose profiles of open fields were measured using EBT3 films and compared with the data obtained using an ionization chamber. Eighteen IMRT plans with different field sizes were measured and verified with EBT3 films, applying our methods, and compared to TPS dose maps, to check correct implementation of film dosimetry proposed here. Results The uncertainty of lateral effects can be reduced to ±1 cGy. Compared with the results of Micke A et al., the residual disturbances of the proposed triple-channel method at 48, 176 and 415 cGy are 5.3%, 20.9% and 31.4% smaller, respectively. Compared with the ionization chamber results, the difference in the off-axis ratio and percentage depth dose are within 1% and 2%, respectively. For the application of IMRT verification, there were no difference between two triple-channel methods. Compared with only corrected by triple-channel method, the IMRT results of the combined method (include lateral effect correction and our present triple-channel method) show a 2% improvement for large IMRT fields with the criteria 3%/3 mm. PMID:28750023

  11. SU-E-T-76: Comparing Homogeneity Between Gafchromic Film EBT2 and EBT3

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Mizuno, H; Sumida, I; Ogawa, K

    2014-06-01

    Purpose: We found out that homogeneity of EBT2 was different among lot numbers in previous study. Variation in local homogeneity of EBT3 among several lot numbers has not been reported. In this study, we investigated film homogeneity of Gafcrhomic EBT3 films compared with EBT2 films. Methods: All sheets from five lots were cut into 12 pieces to investigate film homogeneity, and were irradiated at 0.5, 2, and 3 Gy. To investigate intra- and inter-sheet uniformity, five sheets from five lots were exposed to 2 Gy: intra-sheet uniformity was evaluated by the coefficient of variation of homogeneity for all pieces ofmore » a single sheet, and inter-sheet uniformity was evaluated by the coefficient of variation of homogeneity among the same piece numbers in the five sheets. To investigate the difference of ADC value in various doses, a single sheet from each of five lots was irradiated at 0.5 Gy and 3 Gy in addition to 2 Gy. A scan resolution of 72 dots per inch (dpi) and color depth of 48-bit RGB were used. Films were analyzed by the inhouse software; Average of ADC value in center ROI and profile X and Y axis were measured. Results and Conclusion: Intra-sheet uniformity of non-irradiated EBT2 films were ranged from 0.1% to 0.4%, however that of irradiated EBT2 films were ranged from 0.2% to 1.5%. On the other hand, intra-sheet uniformity of irradiated and non-irradiated EBT3 films were from 0.2% to 0.6%. Inter-sheet uniformity of all films were less than 0.5%. It was interesting point that homogeneity of EBT3 between no-irradiated and irradiated films were similar value, whereas EBT2 had dose dependence of homogeneity in ADC value evaluation. These results suggested that EBT3 homogeneity was corrected by this feature.« less

  12. Characterizing the marker-dye correction for Gafchromic(®) EBT2 film: a comparison of three analysis methods.

    PubMed

    McCaw, Travis J; Micka, John A; Dewerd, Larry A

    2011-10-01

    Gafchromic(®) EBT2 film has a yellow marker dye incorporated into the active layer of the film that can be used to correct the film response for small variations in thickness. This work characterizes the effect of the marker-dye correction on the uniformity and uncertainty of dose measurements with EBT2 film. The effect of variations in time postexposure on the uniformity of EBT2 is also investigated. EBT2 films were used to measure the flatness of a (60)Co field to provide a high-spatial resolution evaluation of the film uniformity. As a reference, the flatness of the (60)Co field was also measured with Kodak EDR2 films. The EBT2 films were digitized with a flatbed document scanner 24, 48, and 72 h postexposure, and the images were analyzed using three methods: (1) the manufacturer-recommended marker-dye correction, (2) an in-house marker-dye correction, and (3) a net optical density (OD) measurement in the red color channel. The field flatness was calculated from orthogonal profiles through the center of the field using each analysis method, and the results were compared with the EDR2 measurements. Uncertainty was propagated through a dose calculation for each analysis method. The change in the measured field flatness for increasing times postexposure was also determined. Both marker-dye correction methods improved the field flatness measured with EBT2 film relative to the net OD method, with a maximum improvement of 1% using the manufacturer-recommended correction. However, the manufacturer-recommended correction also resulted in a dose uncertainty an order of magnitude greater than the other two methods. The in-house marker-dye correction lowered the dose uncertainty relative to the net OD method. The measured field flatness did not exhibit any unidirectional change with increasing time postexposure and showed a maximum change of 0.3%. The marker dye in EBT2 can be used to improve the response uniformity of the film. Depending on the film analysis method used

  13. Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films.

    PubMed

    Dreindl, Ralf; Georg, Dietmar; Stock, Markus

    2014-05-01

    Gafchromic® EBT2 film is a widely used dosimetric tool for quality assurance in radiation therapy. In 2012 EBT3 was presented as a replacement for EBT2 films. The symmetric structure of EBT3 films to reduce face-up/down dependency as well as the inclusion of a matte film surface to frustrate Newton Ring artifacts present the most prominent improvements of EBT3 films. The aim of this study was to investigate the characteristics of EBT3 films, to benchmark the films against the known EBT2-features and to evaluate the dosimetric behavior over a time period greater than 6 months. All films were irradiated to clinical photon beams (6 MV, 10 MV and 18 MV) on an Elekta Synergy Linac equipped with a Beam Modulator MLC in solid water phantom slabs. Film digitalization was done with a flatbed transparency scanner (Type Epson Expression 1680 Pro). MATLAB® was used for further statistical calculations and image processing. The investigations on post-irradiation darkening, film orientation, film uniformity and energy dependency resulted in negligible differences between EBT2 and EBT3 film. A minimal improvement in face-up/down dependence was found for EBT3. The matte film surface of EBT3 films turned out to be a practical feature as Newton rings could be eliminated completely. Considering long-term behavior (> 6 months) a shift of the calibration curve for EBT2 and EBT3 films due to changes in the dynamic response of the active component was observed. In conclusion, the new EBT3 film yields comparable results to its predecessor EBT2. The general advantages of radiochromic film dosimeters are completed by high film homogeneity, low energy dependence for the observed energy range and a minimized face-up/down dependence. EBT2 dosimetry-protocols can also be used for EBT3 films, but the inclusion of periodical recalibration-interval (e.g. once a quarter) is recommended for protocols of both film generations. Copyright © 2013. Published by Elsevier GmbH.

  14. The use of new GAFCHROMIC EBT film for 125I seed dosimetry in Solid Water phantom.

    PubMed

    Chiu-Tsao, Sou-Tung; Medich, David; Munro, John

    2008-08-01

    Radiochromic film dosimetry has been extensively used for intravascular brachytherapy applications for near field within 1 cm from the sources. With the recent introduction of new model of radiochromic films, GAFCHROMIC EBT, with higher sensitivity than earlier models, it is promising to extend the distances out to 5 cm for low dose rate (LDR) source dosimetry. In this study, the use of new model GAFCHROMIC EBT film for 125I seed dosimetry in Solid Water was evaluated for radial distances from 0.06 cm out to 5 cm. A multiple film technique was employed for four 125I seeds (Implant Sciences model 3500) with NIST traceable air kerma strengths. Each experimental film was positioned in contact with a 125I seed in a Solid Water phantom. The products of the air kerma strength and exposure time ranged from 8 to 3158 U-h, with the initial air kerma strength of 6 U in a series of 25 experiments. A set of 25 calibration films each was sequentially exposed to one 125I seed at about 0.58 cm distance for doses from 0.1 to 33 Gy. A CCD camera based microdensitometer, with interchangeable green (520 nm) and red (665 nm) light boxes, was used to scan all the films with 0.2 mm pixel resolution. The dose to each 125I calibration film center was calculated using the air kerma strength of the seed (incorporating decay), exposure time, distance from seed center to film center, and TG43U1S1 recommended dosimetric parameters. Based on the established calibration curve, dose conversion from net optical density was achieved for each light source. The dose rate constant was determined as 0.991 cGy U(-1)h(-1) (+/-6.9%) and 1.014 cGy U(-1)h(-1) (+/-6.8%) from films scanned using green and red light sources, respectively. The difference between these two values was within the uncertainty of the measurement. Radial dose function and 2D anisotropy function were also determined. The results obtained using the two light sources corroborated each other. We found good agreement with the TG43U1S1

  15. Dosimetric verification of stereotactic radiosurgery/stereotactic radiotherapy dose distributions using Gafchromic EBT3

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Cusumano, Davide, E-mail: davide.cusumano@unimi.it; Fumagalli, Maria L.; Marchetti, Marcello

    2015-10-01

    Aim of this study is to examine the feasibility of using the new Gafchromic EBT3 film in a high-dose stereotactic radiosurgery and radiotherapy quality assurance procedure. Owing to the reduced dimensions of the involved lesions, the feasibility of scanning plan verification films on the scanner plate area with the best uniformity rather than using a correction mask was evaluated. For this purpose, signal values dispersion and reproducibility of film scans were investigated. Uniformity was then quantified in the selected area and was found to be within 1.5% for doses up to 8 Gy. A high-dose threshold level for analyses usingmore » this procedure was established evaluating the sensitivity of the irradiated films. Sensitivity was found to be of the order of centiGray for doses up to 6.2 Gy and decreasing for higher doses. The obtained results were used to implement a procedure comparing dose distributions delivered with a CyberKnife system to planned ones. The procedure was validated through single beam irradiation on a Gafchromic film. The agreement between dose distributions was then evaluated for 13 patients (brain lesions, 5 Gy/die prescription isodose ~80%) using gamma analysis. Results obtained using Gamma test criteria of 5%/1 mm show a pass rate of 94.3%. Gamma frequency parameters calculation for EBT3 films showed to strongly depend on subtraction of unexposed film pixel values from irradiated ones. In the framework of the described dosimetric procedure, EBT3 films proved to be effective in the verification of high doses delivered to lesions with complex shapes and adjacent to organs at risk.« less

  16. Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification

    PubMed Central

    Borca, Valeria Casanova; Russo, Giuliana; Grosso, Pierangelo; Cante, Domenico; Sciacero, Piera; Girelli, Giuseppe; Porta, Maria Rosa La; Tofani, Santi

    2013-01-01

    Radiochromic film has become an important tool to verify dose distributions in highly conformal radiation therapy such as IMRT. Recently, a new generation of these films, EBT3, has become available. EBT3 has the same composition and thickness of the sensitive layer of the previous EBT2 films, but its symmetric layer configuration allows the user to eliminate side orientation dependence, which is reported for EBT2 films. The most important EBT3 characteristics have been investigated, such as response at high‐dose levels, sensitivity to scanner orientation and postirradiation coloration, energy and dose rate dependence, and orientation dependence with respect to film side. Additionally, different IMRT fields were measured with both EBT3 and EBT2 films and evaluated using gamma index analysis. The results obtained show that most of the characteristics of EBT3 film are similar to the EBT2 film, but the orientation dependence with respect to film side is completely eliminated in EBT3 films. The study confirms that EBT3 film can be used for clinical practice in the same way as the previous EBT2 film. PACS number: 87.56.Fc PMID:23470940

  17. Dosimetric verification of stereotactic radiosurgery/stereotactic radiotherapy dose distributions using Gafchromic EBT3.

    PubMed

    Cusumano, Davide; Fumagalli, Maria L; Marchetti, Marcello; Fariselli, Laura; De Martin, Elena

    2015-01-01

    Aim of this study is to examine the feasibility of using the new Gafchromic EBT3 film in a high-dose stereotactic radiosurgery and radiotherapy quality assurance procedure. Owing to the reduced dimensions of the involved lesions, the feasibility of scanning plan verification films on the scanner plate area with the best uniformity rather than using a correction mask was evaluated. For this purpose, signal values dispersion and reproducibility of film scans were investigated. Uniformity was then quantified in the selected area and was found to be within 1.5% for doses up to 8 Gy. A high-dose threshold level for analyses using this procedure was established evaluating the sensitivity of the irradiated films. Sensitivity was found to be of the order of centiGray for doses up to 6.2 Gy and decreasing for higher doses. The obtained results were used to implement a procedure comparing dose distributions delivered with a CyberKnife system to planned ones. The procedure was validated through single beam irradiation on a Gafchromic film. The agreement between dose distributions was then evaluated for 13 patients (brain lesions, 5 Gy/die prescription isodose ~80%) using gamma analysis. Results obtained using Gamma test criteria of 5%/1 mm show a pass rate of 94.3%. Gamma frequency parameters calculation for EBT3 films showed to strongly depend on subtraction of unexposed film pixel values from irradiated ones. In the framework of the described dosimetric procedure, EBT3 films proved to be effective in the verification of high doses delivered to lesions with complex shapes and adjacent to organs at risk. Copyright © 2015 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

  18. Commissioning Varian enhanced dynamic wedge in the PINNACLE treatment planning system using Gafchromic EBT film.

    PubMed

    Fontanarosa, Davide; Orlandini, Lucia Clara; Andriani, Italo; Bernardi, Luca

    2009-10-01

    In external photon beam therapies, the technique of dynamic wedge is a well established method for dose inhomogeneity compensation. The introduction of the enhanced dynamic wedge (EDW) on Varian LINACs considerably improved the pre-existing techniques. In the process of commissioning a Varian LINAC into a PINNACLE3 treatment planning system (TPS), the user is required to import quite a few measurements of EDW profiles and percentage depth doses (PDDs). Standard measurement devices like ionization chambers in a water phantom are not the most indicated ones for this situation where each measurement point is obtained by integrating during the entire exposure: Measurements would result to be a very laborious and time consuming operation, most of the times not practically possible. The goal of the present work is to introduce an alternative and hands-on procedure to perform the measurements using a combination of GafchromicTM EBT films, irradiated sideways in one single shot for both profiles and PDDs, and a single standard ionization chamber. The scanned profiles obtained at different depths have easily been imported in the TPS; for the PDD measurements, a correction was proven necessary to account for a "self-shielding" effect introduced by the presence of the films themselves, when irradiated sideways, resulting in an underestimation of the dose at deeper depths. A correction curve was derived comparing TPS open field validated measurements with the curves extracted from GafchromicTM EBT films. Finally, the curve was applied to all the wedged fields PDD measurements and could minimize the errors. The comparison for the 15 MV photon beam between the measured and the calculated 48 profiles and 12 PDDs (field sizes from 5 x 5 to 20 x 20 cm2, wedge angles ranging from 15 degrees to 60 degrees) was acceptable. The confidence limit (CL) was used as fit indicator, as suggested by the ESTRO Booklet No. 7: For the investigated PDDs the maximum value was 6.40 in the build up

  19. Homogeneity of GAFCHROMIC EBT2 film among different lot numbers

    PubMed Central

    Takahashi, Yutaka; Tanaka, Atsushi; Hirayama, Takamitsu; Yamaguchi, Tsuyoshi; Katou, Hiroaki; Takahara, Keiko; Okamoto, Yoshiaki; Teshima, Teruki

    2012-01-01

    EBT2 film is widely used for quality assurance in radiation therapy. The purpose of this study was to investigate the homogeneity of EBT2 film among various lots, and the dose dependence of heterogeneity. EBT2 film was positioned in the center of a flatbed scanner and scanned in transmission mode at 75 dpi. Homogeneity was investigated by evaluating gray value and net optical density (netOD) with the red color channel. The dose dependence of heterogeneity in a single sheet from five lots was investigated at 0.5, 2, and 3 Gy. Maximum coefficient of variation as evaluated by netOD in a single film was 3.0% in one lot, but no higher than 0.5% in other lots. Dose dependence of heterogeneity was observed on evaluation by gray value but not on evaluation by netOD. These results suggest that EBT2 should be examined in each lot number before clinical use, and that the dose calibration curve should be constructed using netOD. PACS number: 87 PMID:22766947

  20. A study on rectal dose measurement in phantom and in vivo using Gafchromic EBT3 film in IMRT and CyberKnife treatments of carcinoma of prostate

    PubMed Central

    Ganapathy, K.; Kurup, P. G. G.; Murali, V.; Muthukumaran, M.; Subramanian, S. Balaji; Velmurugan, J.

    2013-01-01

    The objective of this study is to check the feasibility of in vivo rectal dose measurement in intensity-modulated radiotherapy (IMRT) and CyberKnife treatments for carcinoma prostate. An in-house pelvis phantom made with bee's wax was used in this study. Two cylindrical bone equivalent materials were used to simulate the femur. Target and other critical structures associated with carcinoma prostate were delineated on the treatment planning images by the radiation oncologist. IMRT treatment plan was generated in Oncentra Master Plan treatment planning system and CyberKnife treatment plan was generated in Multiplan treatment planning system. Dose measurements were carried out in phantom and in patient using Gafchromic EBT3 films. RIT software was used to analyze the dose measured by EBT3 films. The measured doses using EBT3 films were compared with the TPS-calculated dose along the anterior rectal wall at multiple points. From the in-phantom measurements, it is observed that the difference between calculated and measured dose was mostly within 5%, except for a few measurement points. The difference between calculated and measured dose in the in-patient measurements was higher than 5% in regions which were away from the target. Gafchromic EBT3 film is a suitable detector for in vivo rectal dose measurements as it offers the possibility of analyzing the dose at multiple points. In addition, the method of extending this in vivo rectal dose measurement technique as a tool for patient-specific quality assurance check is also analyzed. PMID:24049320

  1. Simplified method for creating a density-absorbed dose calibration curve for the low dose range from Gafchromic EBT3 film.

    PubMed

    Gotanda, Tatsuhiro; Katsuda, Toshizo; Gotanda, Rumi; Kuwano, Tadao; Akagawa, Takuya; Tanki, Nobuyoshi; Tabuchi, Akihiko; Shimono, Tetsunori; Kawaji, Yasuyuki

    2016-01-01

    Radiochromic film dosimeters have a disadvantage in comparison with an ionization chamber in that the dosimetry process is time-consuming for creating a density-absorbed dose calibration curve. The purpose of this study was the development of a simplified method of creating a density-absorbed dose calibration curve from radiochromic film within a short time. This simplified method was performed using Gafchromic EBT3 film with a low energy dependence and step-shaped Al filter. The simplified method was compared with the standard method. The density-absorbed dose calibration curves created using the simplified and standard methods exhibited approximately similar straight lines, and the gradients of the density-absorbed dose calibration curves were -32.336 and -33.746, respectively. The simplified method can obtain calibration curves within a much shorter time compared to the standard method. It is considered that the simplified method for EBT3 film offers a more time-efficient means of determining the density-absorbed dose calibration curve within a low absorbed dose range such as the diagnostic range.

  2. Simplified method for creating a density-absorbed dose calibration curve for the low dose range from Gafchromic EBT3 film

    PubMed Central

    Gotanda, Tatsuhiro; Katsuda, Toshizo; Gotanda, Rumi; Kuwano, Tadao; Akagawa, Takuya; Tanki, Nobuyoshi; Tabuchi, Akihiko; Shimono, Tetsunori; Kawaji, Yasuyuki

    2016-01-01

    Radiochromic film dosimeters have a disadvantage in comparison with an ionization chamber in that the dosimetry process is time-consuming for creating a density-absorbed dose calibration curve. The purpose of this study was the development of a simplified method of creating a density-absorbed dose calibration curve from radiochromic film within a short time. This simplified method was performed using Gafchromic EBT3 film with a low energy dependence and step-shaped Al filter. The simplified method was compared with the standard method. The density-absorbed dose calibration curves created using the simplified and standard methods exhibited approximately similar straight lines, and the gradients of the density-absorbed dose calibration curves were −32.336 and −33.746, respectively. The simplified method can obtain calibration curves within a much shorter time compared to the standard method. It is considered that the simplified method for EBT3 film offers a more time-efficient means of determining the density-absorbed dose calibration curve within a low absorbed dose range such as the diagnostic range. PMID:28144120

  3. SU-E-T-269: Quality Assurance of Spine Volumetric Modulated Arc Therapy with Flattening Filter Free Beams Using Gafchromic EBT3 Films

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Choi, Y; Cho, B; Kwak, J

    2014-06-01

    Purpose: We implemented the Gafchromic film-based patient specific QA of volumetric modulated arc therapy (VMAT) with flattening-filter free (FFF) beams for spine metastases and validated the accuracy of fast arc delivery. Methods: EBT3 films and a homemade cylindrical QA phantom were employed for dosimetric verification of VMATs. For 14 FFF VMAT plans (10 with 10-MV FFF beams and 4 with 6-MV FFF beams), the doses were recalculated on the phantom and delivered by a TrueBeam STx accelerator equipped with a high-definition 120 leaf MLC. The EBT3 films were scanned using an Epson 10000XL scanner through the FilmQA Pro software. Allmore » the irradiated film images were converted to dose map using a calibration response curve. The resulting dose map of film measurement was compared with treatment plan and evaluated using gamma analysis with dose tolerance of 2% within 2 mm. In addition, the point-dose measurement in the phantom using an ion chamber was evaluated as a reference in a ratio of measured and planned doses. Results: The gamma pass rates averaged over all FFF plans for composite-field measurements were 96.0 ± 3.6% (88.9%–99.5%). When adopting a tolerance level of 3% - 3 mm, the gamma pass rates were improved with the ranges from 98% to 100%. In addition, dose profiles and dose distributions showed that spinal cord was protected by the rapid dose fall-off and by delivering the treatment with high precision. In point-dose measurements, the average differences between the measured and planned doses were 0.5% ± 1.0% of the prescription dose. Conclusion: We demonstrated that Gafchromic EBT3 film would be an effective patient-specific QA tool, especially for VMAT of spine SBRT with treatment of small fields and highly gradient dose distributions. The results of film QA verified that the dosimetric accuracy of spine SBRT utilizing RapidArc with FFF beams in our institution is reliable.« less

  4. Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Brown, Thomas A. D.; Hogstrom, Kenneth R.; Alvarez, Diane

    Purpose: This work investigates the dose-response curves of GAFCHROMIC{sup Registered-Sign} EBT, EBT2, and EBT3 radiochromic films using synchrotron-produced monochromatic x-ray beams. EBT2 film is being utilized for dose verification in photoactivated Auger electron therapy at the Louisiana State University Center for Advanced Microstructures and Devices (CAMD) synchrotron facility. Methods: Monochromatic beams of 25, 30, and 35 keV were generated on the tomography beamline at CAMD. Ion chamber depth-dose measurements were used to determine the dose delivered to films irradiated at depths from 0.7 to 8.5 cm in a 10 Multiplication-Sign 10 Multiplication-Sign 10-cm{sup 3} polymethylmethacrylate phantom. AAPM TG-61 protocol wasmore » applied to convert measured ionization into dose. Films were digitized using an Epson 1680 Professional flatbed scanner and analyzed using the net optical density (NOD) derived from the red channel. A dose-response curve was obtained at 35 keV for EBT film, and at 25, 30, and 35 keV for EBT2 and EBT3 films. Calibrations of films for 4 MV x-rays were obtained for comparison using a radiotherapy accelerator at Mary Bird Perkins Cancer Center. Results: The sensitivity (NOD per unit dose) of EBT film at 35 keV relative to that for 4-MV x-rays was 0.73 and 0.76 for doses 50 and 100 cGy, respectively. The sensitivity of EBT2 film at 25, 30, and 35 keV relative to that for 4-MV x-rays varied from 1.09-1.07, 1.23-1.17, and 1.27-1.19 for doses 50-200 cGy, respectively. For EBT3 film the relative sensitivity was within 3% of unity for all three monochromatic x-ray beams. Conclusions: EBT and EBT2 film sensitivity showed strong energy dependence over an energy range of 25 keV-4 MV, although this dependence becomes weaker for larger doses. EBT3 film shows weak energy dependence, indicating that it would be a better dosimeter for kV x-ray beams where beam hardening effects can result in large changes in the effective energy.« less

  5. Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance.

    PubMed

    Fuss, Martina; Sturtewagen, Eva; De Wagter, Carlos; Georg, Dietmar

    2007-07-21

    The suitability of radiochromic EBT film was studied for high-precision clinical quality assurance (QA) by identifying the dose response for a wide range of irradiation parameters typically modified in highly-conformal treatment techniques. In addition, uncertainties associated with varying irradiation conditions were determined. EBT can be used for dose assessment of absorbed dose levels as well as relative dosimetry when compared to absolute absorbed dose calibrated using ionization chamber results. For comparison, a silver halide film (Kodak EDR-2) representing the current standard in film dosimetry was included. As an initial step a measurement protocol yielding accurate and precise results was established for a flatbed transparency scanner (Epson Expression 1680 Pro) that was utilized as a film reading instrument. The light transmission measured by the scanner was found to depend on the position of the film on the scanner plate. For three film pieces irradiated with doses of 0 Gy, approximately 1 Gy and approximately 7 Gy, the pixel values measured in portrait or landscape mode differed by 4.7%, 6.2% and 10.0%, respectively. A study of 200 film pieces revealed an excellent sheet-to-sheet uniformity. On a long time scale, the optical development of irradiated EBT film consisted of a slow but steady increase of absorbance which was not observed to cease during 4 months. Sensitometric curves of EBT films obtained under reference conditions (SSD = 95 cm, FS = 5 x 5 cm(2), d = 5 cm) for 6, 10 and 25 MV photon beams did not show any energy dependence. The average separation between all curves was only 0.7%. The variation of the depth d (range 2-25 cm) in the phantom did not affect the dose response of EBT film. Also the influence of the radiation field size (range 3 x 3-40 x 40 cm(2)) on the sensitometric curve was not significant. For EDR-2 films maximum differences between the calibration curves reached 7-8% for X6MV and X25MV. Radiochromic EBT film, in

  6. Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance

    NASA Astrophysics Data System (ADS)

    Fuss, Martina; Sturtewagen, Eva; DeWagter, Carlos; Georg, Dietmar

    2007-07-01

    The suitability of radiochromic EBT film was studied for high-precision clinical quality assurance (QA) by identifying the dose response for a wide range of irradiation parameters typically modified in highly-conformal treatment techniques. In addition, uncertainties associated with varying irradiation conditions were determined. EBT can be used for dose assessment of absorbed dose levels as well as relative dosimetry when compared to absolute absorbed dose calibrated using ionization chamber results. For comparison, a silver halide film (Kodak EDR-2) representing the current standard in film dosimetry was included. As an initial step a measurement protocol yielding accurate and precise results was established for a flatbed transparency scanner (Epson Expression 1680 Pro) that was utilized as a film reading instrument. The light transmission measured by the scanner was found to depend on the position of the film on the scanner plate. For three film pieces irradiated with doses of 0 Gy, ~1 Gy and ~7 Gy, the pixel values measured in portrait or landscape mode differed by 4.7%, 6.2% and 10.0%, respectively. A study of 200 film pieces revealed an excellent sheet-to-sheet uniformity. On a long time scale, the optical development of irradiated EBT film consisted of a slow but steady increase of absorbance which was not observed to cease during 4 months. Sensitometric curves of EBT films obtained under reference conditions (SSD = 95 cm, FS = 5 × 5 cm2, d = 5 cm) for 6, 10 and 25 MV photon beams did not show any energy dependence. The average separation between all curves was only 0.7%. The variation of the depth d (range 2-25 cm) in the phantom did not affect the dose response of EBT film. Also the influence of the radiation field size (range 3 × 3-40 × 40 cm2) on the sensitometric curve was not significant. For EDR-2 films maximum differences between the calibration curves reached 7-8% for X6MV and X25MV. Radiochromic EBT film, in combination with a flatbed scanner

  7. SU-F-I-70: Investigation of Gafchromic EBT3 Film Energy Dependence Using Proton, Photon, and Electron Beams

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ferreira, C; Schnell, E; Ahmad, S

    Purpose: To investigate the energy dependence of Gafchromic EBT3 film over a range of clinically used proton, photon and electron energies. Methods: Proton beam energies of 117 and 204 MeV, corresponding respectively to ranges in water of 10 cm and 27 cm from a Mevion S250 double scatter system unit were used. Electron energies of 6 and 20 MeV and photon energies of 6 and 18 MV from a Varian Clinac 21EX Linac were used. Two pieces of film (5×5 cm{sup 2}) were irradiated sequentially for doses of 100, 500, and 1000 cGy for all energies and modalities. Films weremore » placed on the central beam axis for a 10×10 cm{sup 2} field size in the middle of spread out Bragg peak (SOBP) for proton and in respective dmax for photon and electron energies. Films were scanned on a flatbed Epson Expression 10000 XL scanner on the central region of the scanning window using 48-bit, 300 dpi, and landscape orientation after 48 hours post-irradiation of film to account for optical density (OD) stabilization. Film analysis of the red channel was performed using ImageJ 1.48v (National Institutes of Health). Results: The energy dependency of EBT3 among all energies and modalities for all doses studied was small within measurement uncertainties (1σ = ± 4.1%). The mean net OD in red channel for films receiving the same dose in the same energy modality had standard deviations within 0.9% for photons, 4.9% for electrons and 1.8% for protons. It was observed that film pieces were activated during proton irradiation, e.g., 7 mR/hr at surface after 30 minutes of irradiation, lasting for 2 hours post irradiation. Conclusion: EBT3 energy dependency was evaluated for clinically used proton, photon, and electron energies. The film self-activation may have contributed to fog and negligible dose.« less

  8. An experimental extrapolation technique using the Gafchromic EBT3 film for relative output factor measurements in small x-ray fields

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Morales, Johnny E., E-mail: johnny.morales@lh.org.

    Purpose: An experimental extrapolation technique is presented, which can be used to determine the relative output factors for very small x-ray fields using the Gafchromic EBT3 film. Methods: Relative output factors were measured for the Brainlab SRS cones ranging in diameters from 4 to 30 mm{sup 2} on a Novalis Trilogy linear accelerator with 6 MV SRS x-rays. The relative output factor was determined from an experimental reducing circular region of interest (ROI) extrapolation technique developed to remove the effects of volume averaging. This was achieved by scanning the EBT3 film measurements with a high scanning resolution of 1200 dpi.more » From the high resolution scans, the size of the circular regions of interest was varied to produce a plot of relative output factors versus area of analysis. The plot was then extrapolated to zero to determine the relative output factor corresponding to zero volume. Results: Results have shown that for a 4 mm field size, the extrapolated relative output factor was measured as a value of 0.651 ± 0.018 as compared to 0.639 ± 0.019 and 0.633 ± 0.021 for 0.5 and 1.0 mm diameter of analysis values, respectively. This showed a change in the relative output factors of 1.8% and 2.8% at these comparative regions of interest sizes. In comparison, the 25 mm cone had negligible differences in the measured output factor between zero extrapolation, 0.5 and 1.0 mm diameter ROIs, respectively. Conclusions: This work shows that for very small fields such as 4.0 mm cone sizes, a measureable difference can be seen in the relative output factor based on the circular ROI and the size of the area of analysis using radiochromic film dosimetry. The authors recommend to scan the Gafchromic EBT3 film at a resolution of 1200 dpi for cone sizes less than 7.5 mm and to utilize an extrapolation technique for the output factor measurements of very small field dosimetry.« less

  9. Investigation of absorption spectra of Gafchromic EBT2 film's components and their impact on UVR dosimetry

    NASA Astrophysics Data System (ADS)

    Aydarous, Abdulkadir

    2016-05-01

    The absorption spectra of the EBT2 film's components were investigated in conjunction with its use for UVA dosimetry. The polyester (topside) and adhesive layers of the EBT2 film have been gently removed. Gafchromic™ EBT2 films with and without the protected layers (polyester and adhesive) were exposed to UVR of 365 nm for different durations. Thereafter, the UV-visible spectra were measured using a UV-visible spectrophotometer (Model Spectro Dual Split Beam, UVS-2700). Films were digitized using a Nikon CanoScan 9000F Mark II flatbed scanner. The dosimetric characteristics including film's uniformity, reproducibility and post-irradiation development were investigated. The color development of EBT2 and new modified EBT2 (EBT2-M) films irradiated with UVA was relatively stable (less than 1%) immediately after exposure. Based on this study, the sensitivity of EBT2 to UVR with wavelength between ~350 nm and ~390 nm can significantly be enhanced if the adhesive layer (~25 μm) is removed. The polyester layer plays almost no part on absorbing UVR with wavelength between ~320 nm and ~390 nm. Furthermore, various sensitivities for the EBT2-M film has been established depending on the wavelength of analysis.

  10. Measurement of low‐energy backscatter factors using GAFCHROMIC film and OSLDs

    PubMed Central

    Elson, Howard R.; Lamba, Michael A. S.

    2012-01-01

    Some of the lowest voltages used in radiotherapy are termed Grenz and superficial X‐rays of ~ 20 and ~ 100 kVp, respectively. Dosimetrically, the surface doses from these beams are calculated with the use of a free in‐air air kerma measurement combined with a backscatter factor and the appropriate ratio of mass energy absorption coefficients from the measurement material to water. Alternative tools to the standard ion chamber for measuring the BSF are GAFCHROMIC EBT2 film and optically stimulated luminescent dosimeter (OSLD) crystals made from Al2O3. The scope of this project included making three different backscatter measurements with an Xstrahl‐D3100 X‐ray unit on the Grenz ray and superficial settings. These measurements were with OSLDs, GAFCHROMIC EBT2 film, and a PTW ionization chamber. The varied measurement methods allowed for intercomparison to determine the accuracy of the results. The ion chamber measurement was the least accurate, as expected from previous experimental findings. GAFCHROMIC EBT2 film proved to be a useful tool which gave reasonable results, and Landauer OSLDs showed good results for smaller field sizes and an increasing overresponse with larger fields. The specific backscatter factors for this machine demonstrated values about 5% higher than the universal values suggested by the AAPM and IPEMB codes of practice for the 100 kVp setting. The 20 kvp measured data from both techniques showed general agreement with those found in the BJR Supplement No. 10, indicating that this unit's Grenz ray spectrum is similar to those used in previous experimental work. PACS number: 87.53.Bn PMID:23149776

  11. In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment.

    PubMed

    Severgnini, Mara; de Denaro, Mario; Bortul, Marina; Vidali, Cristiana; Beorchia, Aulo

    2014-01-08

    Intraoperative electron radiation therapy (IOERT) cannot usually benefit, as conventional external radiotherapy, from software systems of treatment planning based on computed tomography and from common dose verify procedures. For this reason, in vivo film dosimetry (IVFD) proves to be an effective methodology to evaluate the actual radiation dose delivered to the target. A practical method for IVFD during breast IOERT was carried out to improve information on the dose actually delivered to the tumor target and on the alignment of the shielding disk with respect to the electron beam. Two EBT3 GAFCHROMIC films have been positioned on the two sides of the shielding disk in order to obtain the dose maps at the target and beyond the disk. Moreover the postprocessing analysis of the dose distribution measured on the films provides a quantitative estimate of the misalignment between the collimator and the disk. EBT3 radiochromic films have been demonstrated to be suitable dosimeters for IVD due to their linear dose-optical density response in a narrow range around the prescribed dose, as well as their capability to be fixed to the shielding disk without giving any distortion in the dose distribution. Off-line analysis of the radiochromic film allowed absolute dose measurements and this is indeed a very important verification of the correct exposure to the target organ, as well as an estimate of the dose to the healthy tissue underlying the shielding. These dose maps allow surgeons and radiation oncologists to take advantage of qualitative and quantitative feedback for setting more accurate treatment strategies and further optimized procedures. The proper alignment using elastic bands has improved the absolute dose accuracy and the collimator disk alignment by more than 50%.

  12. In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment

    PubMed Central

    de Denaro, Mario; Bortul, Marina; Vidali, Cristiana; Beorchia, Aulo

    2014-01-01

    Intraoperative electron radiation therapy (IOERT) cannot usually benefit, as conventional external radiotherapy, from software systems of treatment planning based on computed tomography and from common dose verify procedures. For this reason, in vivo film dosimetry (IVFD) proves to be an effective methodology to evaluate the actual radiation dose delivered to the target. A practical method for IVFD during breast IOERT was carried out to improve information on the dose actually delivered to the tumor target and on the alignment of the shielding disk with respect to the electron beam. Two EBT3 GAFCHROMIC films have been positioned on the two sides of the shielding disk in order to obtain the dose maps at the target and beyond the disk. Moreover the postprocessing analysis of the dose distribution measured on the films provides a quantitative estimate of the misalignment between the collimator and the disk. EBT3 radiochromic films have been demonstrated to be suitable dosimeters for IVD due to their linear dose‐optical density response in a narrow range around the prescribed dose, as well as their capability to be fixed to the shielding disk without giving any distortion in the dose distribution. Off‐line analysis of the radiochromic film allowed absolute dose measurements and this is indeed a very important verification of the correct exposure to the target organ, as well as an estimate of the dose to the healthy tissue underlying the shielding. These dose maps allow surgeons and radiation oncologists to take advantage of qualitative and quantitative feedback for setting more accurate treatment strategies and further optimized procedures. The proper alignment using elastic bands has improved the absolute dose accuracy and the collimator disk alignment by more than 50%. PACS number: 87.55.kh

  13. Calibration of GafChromic EBT3 for absorbed dose measurements in 5 MeV proton beam and (60)Co γ-rays.

    PubMed

    Vadrucci, M; Esposito, G; Ronsivalle, C; Cherubini, R; Marracino, F; Montereali, R M; Picardi, L; Piccinini, M; Pimpinella, M; Vincenti, M A; De Angelis, C

    2015-08-01

    To study EBT3 GafChromic film in low-energy protons, and for comparison purposes, in a reference (60)Co beam in order to use it as a calibrated dosimetry system in the proton irradiation facility under construction within the framework of the Oncological Therapy with Protons (TOP)-Intensity Modulated Proton Linear Accelerator for RadioTherapy (IMPLART) Project at ENEA-Frascati, Italy. EBT3 film samples were irradiated at the Istituto Nazionale di Fisica Nucleare-Laboratori Nazionali di Legnaro, Italy, with a 5 MeV proton beam generated by a 7 MV Van de Graaff CN accelerator. The nominal dose rates used were 2.1 Gy/min and 40 Gy/min. The delivered dose was determined by measuring the particle fluence and the energy spectrum in air with silicon surface barrier detector monitors. A preliminary study of the EBT3 film beam quality dependence in low-energy protons was conducted by passively degrading the beam energy. EBT3 films were also irradiated at ENEA-National Institute of Ionizing Radiation Metrology with gamma radiation produced by a (60)Co source characterized by an absorbed dose to water rate of 0.26 Gy/min as measured by a calibrated Farmer type ionization chamber. EBT3 film calibration curves were determined by means of a set of 40 film pieces irradiated to various doses ranging from 0.5 Gy to 30 Gy absorbed dose to water. An EPSON Expression 11000XL color scanner in transmission mode was used for film analysis. Scanner response stability, intrafilm uniformity, and interfilm reproducibility were verified. Optical absorption spectra measurements were performed on unirradiated and irradiated EBT3 films to choose the most sensitive color channel to the dose range used. EBT3 GafChromic films show an under response up to about 33% for low-energy protons with respect to (60)Co gamma radiation, which is consistent with the linear energy transfer dependence already observed with higher energy protons, and a negligible dose-rate dependence in the 2-40 Gy/min range

  14. IMRT plan verification with EBT2 and EBT3 films compared to PTW 2D-ARRAY seven29

    NASA Astrophysics Data System (ADS)

    Hanušová, Tereza; Horáková, Ivana; Koniarová, Irena

    2017-11-01

    The aim of this study was to compare dosimetry with Gafchromic EBT2 and EBT3 films to the ion chamber array PTW seven29 in terms of their performance in clinical IMRT plan verification. A methodology for film processing and calibration was developed. Calibration curves were obtained in MATLAB and in FilmQA Pro. The best calibration curve was then used to calibrate EBT2 and EBT3 films for IMRT plan verification measurements. Films were placed in several coronal planes into an RW3 slab phantom and irradiated with a clinical IMRT plan for prostate and lymph nodes using 18 MV photon beams. Individual fields were tested and irradiated with gantry at 0°. Results were evaluated using gamma analysis with 3%/3 mm criteria in OmniPro I'mRT version 1.7. The same measurements were performed with the ion chamber array PTW seven29 in RW3 slabs (different depths) and in the OCTAVIUS II phantom (isocenter depth only; both original and nominal gantry angles). Results were evaluated in PTW VeriSoft version 3.1 using the same criteria. Altogether, 45 IMRT planes were tested with film and 25 planes with the PTW 2D-ARRAY seven29. Film measuerements showed different results than ion chamber matrix measurements. With PTW 2D-ARRAY seven29, worse results were obtained when the detector was placed into the OCTAVIUS phantom than into the RW3 slab phantom, and the worst pass rates were seen for rotational measurements. EBT2 films showed inconsistent results and could differ significantly for different planes in one field. EBT3 films seemed to give the best results of all the tested configurations.

  15. SU-E-T-637: Age and Batch Dependence of Gafchromic EBT Films in Photon and Proton Beam Dosimetry

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Das, I; Akino, Y

    2014-06-15

    Purpose: Gafchrmoic films have undergone significant changes in characteristic over time reflected by HS, EBT, EBT2, EBT3 name. Interand intra- EBT film variability have been studied and found to be significant. However, age and lot/batch type have not been studied in various radiation beams that are investigated in this study. Methods: Thirteen sets of films; 2 EBT, 6 EBT2 and 5 EBT3 films with different lot number and expiration date were acquired. Films were cut longitudinally in 3 cm width and sandwiched between two solid water slabs that were placed in a water phantom to eliminate air gap. Each setmore » of films were irradiated longitudinally at dmax with 6 and 15 MV photon beams as well as in reference condition (16 cm range, 10 cm SOBP) in our uniform scanning proton beam. Films were scanned using an Epson flatbed scanner (ES-10000G) after 48 hours to achieve full polymerization. The profiles were compared with the depth-dose measured with ionization chamber and net optical density (net OD) were calculated. Results: The net OD versus dose for EBT, EBT2 and EBT3 films of different age showed similar trend but with different slope. Even after calibration, differences are clearly visible in net OD in proton and photon beams. A net OD difference of nearly 0.5 is observed in photon but this was limited to 0.2–0.3 in proton beam. This relates to 20% and 15% dosimetric difference in photon and proton beam respectively over age and type of film. Conclusion: Net OD related to dose is dependent on the age and lot of the film in both photon and proton beams. It is concluded that before any set of film is used, a calibration film should be used for a meaningful dosimetry. The expired films showed larger OD variation compared to unexpired films.« less

  16. Semi-3D dosimetry of high dose rate brachytherapy using a novel Gafchromic EBT3 film-array water phantom

    NASA Astrophysics Data System (ADS)

    Palmer, A. L.; Nisbet, A.; Bradley, D. A.

    2013-06-01

    There is a need to modernise clinical brachytherapy dosimetry measurement beyond traditional point dose verification to enable appropriate quality control within 3D treatment environments. This is to keep pace with the 3D clinical and planning approaches which often include significant patient-specific optimisation away from 'standard loading patterns'. A multi-dimension measurement system is required to provide assurance of the complex 3D dose distributions, to verify equipment performance, and to enable quality audits. However, true 3D dose measurements around brachytherapy applicators are often impractical due to their complex shapes and the requirement for close measurement distances. A solution utilising an array of radiochromic film (Gafchromic EBT3) positioned within a water filled phantom is presented. A calibration function for the film has been determined over 0 to 90Gy dose range using three colour channel analysis (FilmQAPro software). Film measurements of the radial dose from a single HDR source agree with TPS and Monte Carlo calculations within 5 % up to 50 mm from the source. Film array measurements of the dose distribution around a cervix applicator agree with TPS calculations generally within 4 mm distance to agreement. The feasibility of film array measurements for semi-3D dosimetry in clinical HDR applications is demonstrated.

  17. Analysis of the EBT3 Gafchromic film irradiated with 6 MV photons and 6 MeV electrons using reflective mode scanners.

    PubMed

    Farah, Nicolas; Francis, Ziad; Abboud, Marie

    2014-09-01

    We explore in our study the effects of electrons and X-rays irradiations on the newest version of the Gafchromic EBT3 film. Experiments are performed using the Varian "TrueBeam 1.6" medical accelerator delivering 6 MV X-ray photons and 6 MeV electron beams as desired. The main interest is to compare the responses of EBT3 films exposed to two separate beams of electrons and photons, for radiation doses ranging up to 500 cGy. The analysis is done on a flatbed EPSON 10000 XL scanner and cross checked on a HP Scanjet 4850 scanner. Both scanners are used in reflection mode taking into account landscape and portrait scanning positions. After thorough verifications, the reflective scanning method can be used on EBT3 as an economic alternative to the transmission method which was also one of the goals of this study. A comparison is also done between single scan configuration including all samples in a single A4 (HP) or A3 (EPSON) format area and multiple scan procedure where each sample is scanned separately on its own. The images analyses are done using the ImageJ software. Results show significant influence of the scanning configuration but no significant differences between electron and photon irradiations for both single and multiple scan configurations. In conclusion, the film provides a reliable relative dose measurement method for electrons and photons irradiations in the medical field applications. Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  18. GafChromic EBT film dosimetry with flatbed CCD scanner: a novel background correction method and full dose uncertainty analysis.

    PubMed

    Saur, Sigrun; Frengen, Jomar

    2008-07-01

    Film dosimetry using radiochromic EBT film in combination with a flatbed charge coupled device scanner is a useful method both for two-dimensional verification of intensity-modulated radiation treatment plans and for general quality assurance of treatment planning systems and linear accelerators. Unfortunately, the response over the scanner area is nonuniform, and when not corrected for, this results in a systematic error in the measured dose which is both dose and position dependent. In this study a novel method for background correction is presented. The method is based on the subtraction of a correction matrix, a matrix that is based on scans of films that are irradiated to nine dose levels in the range 0.08-2.93 Gy. Because the response of the film is dependent on the film's orientation with respect to the scanner, correction matrices for both landscape oriented and portrait oriented scans were made. In addition to the background correction method, a full dose uncertainty analysis of the film dosimetry procedure was performed. This analysis takes into account the fit uncertainty of the calibration curve, the variation in response for different film sheets, the nonuniformity after background correction, and the noise in the scanned films. The film analysis was performed for film pieces of size 16 x 16 cm, all with the same lot number, and all irradiations were done perpendicular onto the films. The results show that the 2-sigma dose uncertainty at 2 Gy is about 5% and 3.5% for landscape and portrait scans, respectively. The uncertainty gradually increases as the dose decreases, but at 1 Gy the 2-sigma dose uncertainty is still as good as 6% and 4% for landscape and portrait scans, respectively. The study shows that film dosimetry using GafChromic EBT film, an Epson Expression 1680 Professional scanner and a dedicated background correction technique gives precise and accurate results. For the purpose of dosimetric verification, the calculated dose distribution

  19. SU-E-T-475: Improvements to Total Body Irradiation Dosimetry Efficiency with EBT3 Radiochromic Film and a Template System

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Butson, M; Pope, D; Whitaker, M

    Purpose: Total Body Irradiation (TBI) treatments are mainly used in a preparative regimen for haematopoietic stem cell (or bone marrow) transplantation. Our standard regimen is a 12 Gy / 6 fraction bi-daily technique. To evaluate the delivered dose homogeneity to the patient, EBT3 Gafchromic film is positioned at the head, neck, chest, pelvis and groin for all fractions. A system has been developed to simply and accurately prepare and readout the films for patient dose assessment. Methods: A process involving easy preparation and analysis has been produced to minimise the time requirements for TBI dosimetry. One sheet of EBT3 filmmore » is used to prepare treatment dosimeters for all fractions, including calibration films, and an automated dose analysis system for easy evaluation and calculation of estimated in-vivo doses was developed. A desktop scanner is used with a dedicated TBI film template to accurately position the films for Image J analysis and extraction. Dental wax bolus and zip-lock bag holders are used to hold the EBT3 film in place during irradiation. Results: To adequately provide dosimetry information for a 6 fraction, TBI patient, only one sheet of Gafchromic EBT3 film is required. The dosimeters are cut, using a template, into 19 mm squares which are then placed between two 30 mm x 30 mm x 4.5 mm wax blocks for bolus. All packages are prepared before the first treatment fraction. The scanning and analysis process can be completed in less than 10 minutes after a 240 min development period. Results have shown that a high level of accuracy and reproducibility can be achieved using the template system provided. Conclusion: Gafchromic EBT3 film provides an adequate in-vivo dosimetry measure for TBI patients. Using a template based system on a dedicated desktop scanner, in-vivo results can be ascertained quickly and accurately.« less

  20. From the limits of the classical model of sensitometric curves to a realistic model based on the percolation theory for GafChromic EBT films.

    PubMed

    del Moral, F; Vázquez, J A; Ferrero, J J; Willisch, P; Ramírez, R D; Teijeiro, A; López Medina, A; Andrade, B; Vázquez, J; Salvador, F; Medal, D; Salgado, M; Muñoz, V

    2009-09-01

    Modern radiotherapy uses complex treatments that necessitate more complex quality assurance procedures. As a continuous medium, GafChromic EBT films offer suitable features for such verification. However, its sensitometric curve is not fully understood in terms of classical theoretical models. In fact, measured optical densities and those predicted by the classical models differ significantly. This difference increases systematically with wider dose ranges. Thus, achieving the accuracy required for intensity-modulated radiotherapy (IMRT) by classical methods is not possible, plecluding their use. As a result, experimental parametrizations, such as polynomial fits, are replacing phenomenological expressions in modern investigations. This article focuses on identifying new theoretical ways to describe sensitometric curves and on evaluating the quality of fit for experimental data based on four proposed models. A whole mathematical formalism starting with a geometrical version of the classical theory is used to develop new expressions for the sensitometric curves. General results from the percolation theory are also used. A flat-bed-scanner-based method was chosen for the film analysis. Different tests were performed, such as consistency of the numeric results for the proposed model and double examination using data from independent researchers. Results show that the percolation-theory-based model provides the best theoretical explanation for the sensitometric behavior of GafChromic films. The different sizes of active centers or monomer crystals of the film are the basis of this model, allowing acquisition of information about the internal structure of the films. Values for the mean size of the active centers were obtained in accordance with technical specifications. In this model, the dynamics of the interaction between the active centers of GafChromic film and radiation is also characterized by means of its interaction cross-section value. The percolation model

  1. Surface dose measurement with Gafchromic EBT3 film for intensity modulated radiotherapy technique

    NASA Astrophysics Data System (ADS)

    Akbas, Ugur; Kesen, Nazmiye Donmez; Koksal, Canan; Okutan, Murat; Demir, Bayram; Becerir, Hatice Bilge

    2017-09-01

    Accurate dose measurement in the buildup region is extremely difficult. Studies have reported that treatment planning systems (TPS) cannot calculate surface dose accurately. The aim of the study was to compare the film measurements and TPS calculations for surface dose in head and neck cancer treatment using intensity modulated radiation therapy (IMRT). IMRT plans were generated for 5 head and neck cancer patients by using Varian Eclipse TPS. Quality assurance (QA) plans of these IMRT plans were created on rando phantoms for surface dose measurements. EBT3 films were cut in size of 2.5 x 2.5 cm2 and placed on the left side, right side and the center of larynx and then the films were irradiated with 6 MV photon beams. The measured doses were compared with TPS. The results of TPS calculations were found to be lower compared to the EBT3 film measurements at all selected points. The lack of surface dose calculation in TPS should be considered while evaluating the radiotherapy plans.

  2. Investigation of EBT2 and EBT3 films for proton dosimetry in the 4-20 MeV energy range.

    PubMed

    Reinhardt, S; Würl, M; Greubel, C; Humble, N; Wilkens, J J; Hillbrand, M; Mairani, A; Assmann, W; Parodi, K

    2015-03-01

    Radiochromic films such as Gafchromic EBT2 or EBT3 films are widely used for dose determination in radiation therapy because they offer a superior spatial resolution compared to any other digital dosimetric 2D detector array. The possibility to detect steep dose gradients is not only attractive for intensity-modulated radiation therapy with photons but also for intensity-modulated proton therapy. Their characteristic dose rate-independent response makes radiochromic films also attractive for dose determination in cell irradiation experiments using laser-driven ion accelerators, which are currently being investigated as future medical ion accelerators. However, when using these films in ion beams, the energy-dependent dose response in the vicinity of the Bragg peak has to be considered. In this work, the response of these films for low-energy protons is investigated. To allow for reproducible and background-free irradiation conditions, the films were exposed to mono-energetic protons from an electrostatic accelerator, in the 4-20 MeV energy range. For comparison, irradiation with clinical photons was also performed. It turned out that in general, EBT2 and EBT3 films show a comparable performance. For example, dose-response curves for photons and protons with energies as low as 11 MeV show almost no differences. However, corrections are required for proton energies below 11 MeV. Care has to be taken when correction factors are related to an average LET from depth-dose measurements, because only the dose-averaged LET yields similar results as obtained in mono-energetic measurements.

  3. Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapy

    NASA Astrophysics Data System (ADS)

    Richley, L.; John, A. C.; Coomber, H.; Fletcher, S.

    2010-05-01

    A new radiochromic film, the yellow Gafchromic EBT2, has been marketed as a drop-in replacement for the discontinued blue EBT film. In order to verify the manufacturer's claims prior to clinical use, EBT2 was characterized in transmission, and the less commonly used, reflection modes with an Epson Expression 10000XL A3 flatbed scanner. The red channel was confirmed to provide the greatest sensitivity and was used for all measurements. The post-irradiation darkening of the film was investigated, and the relative response was found to be dose dependent with higher doses stabilizing earlier than lower doses. After 13 h all dose levels had stabilized to within 1% of their value at 24 h. Uniformity of irradiated EBT2 films was within 0.8% and 1.2% (2SD of signal), in reflection and transmission modes, respectively. The light scattering effect, arising from the structure and thickness of EBT2, was found to give rise to an apparent scanner non-uniformity of up to 5.5% in signal. In reflection mode, differences of up to 1.2% were found between the signal obtained from a small film fragment (5 × 5 cm2) and the signal obtained from the same fragment bordered by extra film. Further work is needed to determine the origin of this effect, as there will be implications for reflection dosimetry of intensity modulated fields; reflection mode cannot yet be regarded as a viable alternative to transmission mode. Our results suggest that EBT2 film is a valid alternative, rather than a direct replacement for EBT film.

  4. SU-E-T-118: Dose Verification for Accuboost Applicators Using TLD, Ion Chamber and Gafchromic Film Measurements

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chisela, W; Yao, R; Dorbu, G

    Purpose: To verify dose delivered with HDR Accuboost applicators using TLD, ion chamber and Gafchromic film measurements and to examine applicator leakage. Methods: A microSelectron HDR unit was used to deliver a dose of 50cGy to the mid-plane of a 62mm thick solid water phantom using dwell times from Monte Carlo pre-calculated nomograms for a 60mm, 70mm Round and 60mm Skin-Dose Optimized (SDO) applicators respectively. GafChromic EBT3+ film was embedded in the phantom midplane horizontally to measure dose distribution. Absolute dose was also measured with TLDs and an ADCL calibrated parallel-plate ion chamber placed in the film plane at fieldmore » center for each applicator. The film was calibrated using 6MV x-ray beam. TLDs were calibrated in a Cs-137 source at UW-Madison calibration laboratory. Radiation leakage through the tungsten alloy shell was measured with a film wrapped around outside surface of a 60mm Round applicator. Results: Measured maximum doses at field center are consistently lower than predicated by 5.8% for TLD, 8.8% for ion chamber, and 2.6% for EBT3+ film on average, with measurement uncertainties of 2.2%, 0.3%, and 2.9% for TLD, chamber, film respectively. The total standard uncertainties for ion chamber and Gafchromic film measurement are 4.9% and 4.6% respectively[1]. The area defined by the applicator aperture was covered by 80% of maximum dose for 62mm compression thickness. When 100cGy is delivered to mid-plane with a 60mm Round applicator, surface dose ranges from 60cGy to a maximum of 145cGy, which occurs at source entrance to the applicator. Conclusion: Measured doses by all three techniques are consistently lower than predicted in our measurements. For a compression thickness of 62 mm, the field size defined by the applicator is only covered by 80% of prescribed dose. Radiation leakage of up to 145cGy was found at the source entrance of applicators.« less

  5. Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams

    NASA Astrophysics Data System (ADS)

    Khachonkham, Suphalak; Dreindl, Ralf; Heilemann, Gerd; Lechner, Wolfgang; Fuchs, Hermann; Palmans, Hugo; Georg, Dietmar; Kuess, Peter

    2018-03-01

    Recently, a new type of radiochromic film, the EBT-XD film, has been introduced for high dose radiotherapy. The EBT-XD film contains the same structure as the EBT3 film but has a slightly different composition and a thinner active layer. This study benchmarks the EBT-XD against EBT3 film for 6 MV and 10 MV photon beams, as well as for 97.4 MeV and 148.2 MeV proton beams and 15-100 kV x-rays. Dosimetric and film reading characteristics, such as post irradiation darkening, film orientation effect, lateral response artifact (LRA), film sensitivity, energy and beam quality dependency were investigated. Furthermore, quenching effects in the Bragg peak were investigated for a single proton beam energy for both film types, in addition measurements were performed in a spread-out Bragg peak. EBT-XD films showed the same characteristic on film darkening as EBT3. The effects between portrait and landscape orientation were reduced by 3.1% (in pixel value) for EBT-XD compared to EBT3 at a dose of 2000 cGy. The LRA is reduced for EBT-XD films for all investigated dose ranges. The sensitivity of EBT-XD films is superior to EBT3 for doses higher than 500 cGy. In addition, EBT-XD showed a similar dosimetric response for photon and proton irradiation with low energy and beam quality dependency. A quenching effect of 10% was found for both film types. The slight decrease in the thickness of the active layer and different composition configuration of EBT-XD resulted in a reduced film orientation effect and LRA, as well as a sensitivity increase in high-dose regions for both photon and proton beams. Overall, the EBT-XD film improved regarding film reading characteristics and showed advantages in the high-dose region for photon and proton beams.

  6. Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams.

    PubMed

    Khachonkham, Suphalak; Dreindl, Ralf; Heilemann, Gerd; Lechner, Wolfgang; Fuchs, Hermann; Palmans, Hugo; Georg, Dietmar; Kuess, Peter

    2018-03-15

    Recently, a new type of radiochromic film, the EBT-XD film, has been introduced for high dose radiotherapy. The EBT-XD film contains the same structure as the EBT3 film but has a slightly different composition and a thinner active layer. This study benchmarks the EBT-XD against EBT3 film for 6 MV and 10 MV photon beams, as well as for 97.4 MeV and 148.2 MeV proton beams and 15-100 kV x-rays. Dosimetric and film reading characteristics, such as post irradiation darkening, film orientation effect, lateral response artifact (LRA), film sensitivity, energy and beam quality dependency were investigated. Furthermore, quenching effects in the Bragg peak were investigated for a single proton beam energy for both film types, in addition measurements were performed in a spread-out Bragg peak. EBT-XD films showed the same characteristic on film darkening as EBT3. The effects between portrait and landscape orientation were reduced by 3.1% (in pixel value) for EBT-XD compared to EBT3 at a dose of 2000 cGy. The LRA is reduced for EBT-XD films for all investigated dose ranges. The sensitivity of EBT-XD films is superior to EBT3 for doses higher than 500 cGy. In addition, EBT-XD showed a similar dosimetric response for photon and proton irradiation with low energy and beam quality dependency. A quenching effect of 10% was found for both film types. The slight decrease in the thickness of the active layer and different composition configuration of EBT-XD resulted in a reduced film orientation effect and LRA, as well as a sensitivity increase in high-dose regions for both photon and proton beams. Overall, the EBT-XD film improved regarding film reading characteristics and showed advantages in the high-dose region for photon and proton beams.

  7. Accuracy and efficiency of published film dosimetry techniques using a flat-bed scanner and EBT3 film.

    PubMed

    Spelleken, E; Crowe, S B; Sutherland, B; Challens, C; Kairn, T

    2018-03-01

    Gafchromic EBT3 film is widely used for patient specific quality assurance of complex treatment plans. Film dosimetry techniques commonly involve the use of transmission scanning to produce TIFF files, which are analysed using a non-linear calibration relationship between the dose and red channel net optical density (netOD). Numerous film calibration techniques featured in the literature have not been independently verified or evaluated. A range of previously published film dosimetry techniques were re-evaluated, to identify whether these methods produce better results than the commonly-used non-linear, netOD method. EBT3 film was irradiated at calibration doses between 0 and 4000 cGy and 25 pieces of film were irradiated at 200 cGy to evaluate uniformity. The film was scanned using two different scanners: The Epson Perfection V800 and the Epson Expression 10000XL. Calibration curves, uncertainty in the fit of the curve, overall uncertainty and uniformity were calculated following the methods described by the different calibration techniques. It was found that protocols based on a conventional film dosimetry technique produced results that were accurate and uniform to within 1%, while some of the unconventional techniques produced much higher uncertainties (> 25% for some techniques). Some of the uncommon methods produced reliable results when irradiated to the standard treatment doses (< 400 cGy), however none could be recommended as an efficient or accurate replacement for a common film analysis technique which uses transmission scanning, red colour channel analysis, netOD and a non-linear calibration curve for measuring doses up to 4000 cGy when using EBT3 film.

  8. Characterizing the discoloration of EBT3 films in solar UV A+B measurement using red LED

    NASA Astrophysics Data System (ADS)

    Omar, Ahmad Fairuz; Osman, Ummi Shuhada; Tan, Kok Chooi

    2017-09-01

    This research article proposes an alternative method to measure the discoloration or the color changes of EBT3 films due to exposure by solar ultraviolet (UV A+B) dose. Common methods to measure the color changes of EBT3 are through imaging technique measured by flatbed scanner and through absorbance spectroscopy measured by visible spectrometer. The research presented in this article measure the color changes of EBT3 through simplified optical system using the combination of light emitting diode (LED) as the light source and photodiode as the detector. In this research, 50 pieces of Gafchromic EBT3 films were prepared with the dimension of 3 cm x 2 cm. Color of the films changed from light green to dark green based on the total accumulated UV dose (mJ/cm2) by each film that depends on the duration of exposure, irradiance level (mW/cm2) and condition of the sky. The exposed films were then taken to the laboratory for its color measurement using absorbance spectroscopy technique and using newly developed simplified optical instrument using LED-photodiode. Results from spectroscopy technique indicate that wavelength within red region exhibit better response in term of linearity and responsivity towards the colors of EBT3 films. Wavelength of 626 nm was then selected as the peak emission wavelength for LED-photodiode absorbance system. UV dose measurement using LEDphotodiode system produced good result with coefficient of determination, R2 of 0.97 and root mean square of error, RMSE of 431.82 mJ/cm2 while comparatively, similar wavelength but analyzed from spectroscopy dataset produced R2 of 0.988 and RMSE of 268.94 mJ/cm2.

  9. Evaluation of a novel triple-channel radiochromic film analysis procedure using EBT2.

    PubMed

    van Hoof, Stefan J; Granton, Patrick V; Landry, Guillaume; Podesta, Mark; Verhaegen, Frank

    2012-07-07

    A novel approach to read out radiochromic film was introduced recently by the manufacturer of GafChromic film. In this study, the performance of this triple-channel film dosimetry method was compared against the conventional single-red-channel film dosimetry procedure, with and without inclusion of a pre-irradiation (pre-IR) film scan, using EBT2 film and kilo- and megavoltage photon beams up to 10 Gy. When considering regions of interest averaged doses, the triple-channel method and both single-channel methods produced equivalent results. Absolute dose discrepancies between the triple-channel method, both single-channel methods and the treatment planning system calculated dose values, were no larger than 5 cGy for dose levels up to 2.2 Gy. Signal to noise in triple-channel dose images was found to be similar to signal to noise in single-channel dose images. The accuracy of resulting dose images from the triple- and single-channel methods with inclusion of pre-IR film scan was found to be similar. Results of a comparison of EBT2 data from a kilovoltage depth dose experiment to corresponding Monte Carlo depth dose data produced dose discrepancies of 9.5 ± 12 cGy and 7.6 ± 6 cGy for the single-channel method with inclusion of a pre-IR film scan and the triple-channel method, respectively. EBT2 showed to be energy sensitive at low kilovoltage energies with response differences of 11.9% and 15.6% in the red channel at 2 Gy between 50-225 kVp and 80-225 kVp photon spectra, respectively. We observed that the triple-channel method resulted in non-uniformity corrections of ±1% and consistency values of 0-3 cGy for the batches and dose levels studied. Results of this study indicate that the triple-channel radiochromic film read-out method performs at least as well as the single-channel method with inclusion of a pre-IR film scan, reduces film non-uniformity and saves time with elimination of a pre-IR film scan.

  10. Note: Calibration of EBT3 radiochromic film for measuring solar ultraviolet radiation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chun, S. L.; Yu, P. K. N., E-mail: peter.yu@cityu.edu.hk; State Key Laboratory in Marine Pollution, City University of Hong Kong, Kowloon Tong

    Solar (UVA + UVB) exposure was assessed using the Gafchromic EBT3 film. The coloration change was represented by the net reflective optical density (Net ROD). Through calibrations against a UV-tube lamp, operational relationships were obtained between Net ROD and the (UVA + UVB) exposures (in J cm⁻²p or J m⁻²). The useful range was from ~0.2 to ~30 J cm⁻². The uniformity of UV irradiation was crucial for an accurate calibration. For solar exposures ranging from 2 to 11 J cm⁻², the predicted Net ROD agreed with the recorded values within 9%, while the predicted exposures agreed with the recordedmore » values within 15%.« less

  11. A protocol for EBT3 radiochromic film dosimetry using reflection scanning

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Papaconstadopoulos, Pavlos, E-mail: pavpapac@gmail.com; Hegyi, Gyorgy; Seuntjens, Jan

    2014-12-15

    Purpose: To evaluate the performance of the EBT3 radiochromic film dosimetry system using reflection measurements and to suggest a calibration protocol for precise and accurate reflection film dosimetry. Methods: A set of 14 Gafchromic EBT3 film pieces were irradiated to various doses ranging from 0 to 8 Gy and subsequently scanned using both the reflection and transmission mode. Scanning resolution varied from 50 to 508 dpi (0.5–0.05 mm/pixel). Both the red and green color channels of scanned images were used to relate the film response to the dose. A sensitivity, uncertainty, and accuracy analysis was performed for all scanning modesmore » and color channels. The total uncertainty, along with the fitting and experimental uncertainty components, was identified and analyzed. A microscope resolution target was used to evaluate possible resolution losses under reflection scanning. The calibration range was optimized for reflection scanning in the low (<2 Gy) and high (>2 Gy) dose regions based on the reported results. Results: Reflection scanning using the red channel exhibited the highest sensitivity among all modes, being up to 150% higher than transmission mode in the red channel for the lowest dose level. Furthermore, there was no apparent loss in resolution between the two modes. However, higher uncertainties and reduced accuracy were observed for the red channel under reflection mode, especially at dose levels higher than 2 Gy. These uncertainties were mainly attributed to saturation effects which were translated in poor fitting results. By restricting the calibration to the 0–2 Gy dose range, the situation is reversed and the red reflection mode was superior to the transmission mode. For higher doses, the green channel in reflection mode presented comparable results to the red transmission. Conclusions: A two-color reflection scanning protocol can be suggested for EBT3 radiochromic film dosimetry using the red channel for doses less than 2 Gy and the

  12. Changes of the optical characteristics of radiochromic films in the transition from EBT3 to EBT-XD films.

    PubMed

    Schoenfeld, Andreas A; Wieker, Soeren; Harder, Dietrich; Poppe, Bjoern

    2016-07-01

    A new type of radiochromic film, the EBT-XD film, has been introduced with the aim to reduce the orientation effect and the lateral response artifact occurring in the use of radiochromic films together with flatbed scanners. The task of the present study is to quantify the changes of optical characteristics involved with the transition from the well-known EBT3 films to the new EBT-XD films, using the optical bench arrangement already applied by Schoenfeld et al (2014 Phys. Med. Biol. 59 3575-97). Largely reduced polarization effects and the almost complete loss of the anisotropy of the scattered light produced in a radiation-exposed film have been observed. The Rayleigh-Debye-Gans theory is used to understand these optical changes as arising from the reduced length-to-width ratio of the LiPCDA polymer crystals in the active layer of the EBT-XD film. The effect of these changes on the flatbed scanning artifacts will be shortly addressed, but treated in more detail in a further paper.

  13. Changes of the optical characteristics of radiochromic films in the transition from EBT3 to EBT-XD films

    NASA Astrophysics Data System (ADS)

    Schoenfeld, Andreas A.; Wieker, Soeren; Harder, Dietrich; Poppe, Bjoern

    2016-07-01

    A new type of radiochromic film, the EBT-XD film, has been introduced with the aim to reduce the orientation effect and the lateral response artifact occurring in the use of radiochromic films together with flatbed scanners. The task of the present study is to quantify the changes of optical characteristics involved with the transition from the well-known EBT3 films to the new EBT-XD films, using the optical bench arrangement already applied by Schoenfeld et al (2014 Phys. Med. Biol. 59 3575-97). Largely reduced polarization effects and the almost complete loss of the anisotropy of the scattered light produced in a radiation-exposed film have been observed. The Rayleigh-Debye-Gans theory is used to understand these optical changes as arising from the reduced length-to-width ratio of the LiPCDA polymer crystals in the active layer of the EBT-XD film. The effect of these changes on the flatbed scanning artifacts will be shortly addressed, but treated in more detail in a further paper.

  14. SU-E-T-206: Comparison of EBT and EBT3 RadioChromic Films in Radiation Field of Parotid Cancer Radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Toosi, T Bahreyni; Mianaei, F Khorshidi; Ghorbani, M

    Purpose: The aim of the current study is to compare EBT and EBT3 RadioChromic films in dosimetry of radiotherapy fields for treatment of parotid cancer. Methods: The calibrations of EBT and EBT3 films were performed with the same setups for doses ranging from 0.2 Gy to 5 Gy using 6 MV photon beam of a Siemens Primus linac. These films were scanned in color mode (RGB) by a Microtek (1000XL) scanner and the red color channel data was extracted. Treatment planning for parotid cancer radiation therapy was performed on a RANDO phantom. Skin dose was measured at different points inmore » the right anterior oblique (RAO) and right posterior oblique (RPO) fields by EBT and EBT3 films. Results: Dosimetry was performed with the same conditions for the two film types for calibration and in-phantom in parotid cancer radiotherapy. The measured net optical density (NOD) in EBT film was in some extent higher than that from EBT3 film. The minimum difference between these two films under calibration conditions was about 2.9% (for 0.2 Gy). However, the maximum difference was 35.5% (for 0.5 Gy). In the therapeutic fields of parotid cancer radiotherapy at different points, the measured dose from EBT film was higher than the EBT3 film. In these fields the minimum and maximum measured dose differences were 16.0% and 25.5%, respectively. Conclusion: With the same irradiation and reading conditions, EBT film demonstrates higher NOD than the EBT3 film. This effect may be related to the higher sensitivity of EBT film over EBT3 film. However, the obtained dose differences between these two films in low dose range can be due to the differences in fitting functions applied following the calibration process.« less

  15. Multi-dimensional dosimetric verification of stereotactic radiotherapy for uveal melanoma using radiochromic EBT film.

    PubMed

    Sturtewagen, Eva; Fuss, Martina; Paelinck, Leen; De Wagter, Carlos; Georg, Dietmar

    2008-01-01

    Since 1997, linac based stereotactic radiotherapy (SRT) of uveal melanoma has been continuously developed at the Department of Radiotherapy, Medical University Vienna. The aim of the present study was (i) to test a new type of radiochromic film (Gafchromic EBT) for dosimetric verification of class solutions for these treatments and (ii) to verify treatment plan acceptance criteria, which are based on gamma values statisitcs. An EPSON Expression 1680 Pro flat bed scanner was utilized for film reading. To establish a calibration curve, films were cut in squares of 2 x 2 cm2, positioned at 5 cm depth in a solid water phantom and were irradiated with different dose levels (0.5 and 5 Gy) in a 5 x 5 cm2 field at 6 MV. A previously developed solid phantom (polystyrene) was used with overall dimensions corresponding to an average human head. EBT films were placed at four different depths (10, 20, 25 and 30 mm) and all films were irradiated simultaneously. Four different treatment plans were verified that resemble typical clinical situations. These plans differed in irradiation technique (conformal mMLC or circular arc SRT) and in tumour size (PTV of 1 or 2.5 cm3). In-house developed software was applied to calculate gamma (gamma) index values and to perform several statistical operations (e.g. gamma-area histograms). At depths of 10 mm gamma1%, (gamma-value where 1% of the points have an equal or higher value in the region of interest) were between 1-3 and maximum gamma > 1 (% of gamma-values > 1 in the region of interest) areas were almost 30%. At larger depths, i.e. more close to the isocenter, gamma 1% was < 1 and gamma > 1 areas were mostly < 5%. Average gamma values were about 0.5. Besides the compromised accuracy in the buildup region, previously defined IMRT acceptance criteria [Stock et al., Phys. Med Biol. 50 (2005) 399-411] could be applied as well to SRT. Radiochromic EBT films, in combination with a flat-bed scanner, were found to be an ideal multidimensional

  16. Accurate dosimetry with GafChromic EBT film of a 6 MV photon beam in water: what level is achievable?

    PubMed

    van Battum, L J; Hoffmans, D; Piersma, H; Heukelom, S

    2008-02-01

    This paper focuses on the accuracy, in absolute dose measurements, with GafChromicTM EBT film achievable in water for a 6 MV photon beam up to a dose of 2.3 Gy. Motivation is to get an absolute dose detection system to measure up dose distributions in a (water) phantom, to check dose calculations. An Epson 1680 color (red green blue) transmission flatbed scanner has been used as film scanning system, where the response in the red color channel has been extracted and used for the analyses. The influence of the flatbed film scanner on the film based dose detection process was investigated. The scan procedure has been optimized; i.e. for instance a lateral correction curve was derived to correct the scan value, up to 10%, as a function of optical density and lateral position. Sensitometric curves of different film batches were evaluated in portrait and landscape scan mode. Between various batches important variations in sensitometric curve were observed. Energy dependence of the film is negligible, while a slight variation in dose response is observed for very large angles between film surface and incident photon beam. Improved accuracy in absolute dose detection can be obtained by repetition of a film measurement to tackle at least the inherent presence of film inhomogeneous construction. We state that the overall uncertainty is random in absolute EBT film dose detection and of the order of 1.3% (1 SD) under the condition that the film is scanned in a limited centered area on the scanner and at least two films have been applied. At last we advise to check a new film batch on its characteristics compared to available information, before using that batch for absolute dose measurements.

  17. Ultraviolet exposure of Gafchromic XR‐RV3 and XR‐SP2 films

    PubMed Central

    Gotanda, Rumi; Gotanda, Tatsuhiro; Akagawa, Takuya; Tanki, Nobuyoshi; Kuwano, Tadao; Yabunaka, Kouichi

    2015-01-01

    Gafchromic film has been used for X‐ray dose measurement in diagnostic examinations. Their use has been initiated for three‐dimensional X‐ray dose measurement by using the high‐resolution characteristics of Gafchromic films in computed tomography. However, it is necessary to solve the problem of nonuniform thickness in the active layers of Gafchromic films. A double exposure technique using X‐rays is performed in therapeutic radiology; it is difficult to use in a diagnostic examination because of the heel effect. Therefore, it is suggested that ultraviolet (UV) rays be substituted for X‐rays. However, the appropriate UV wavelength is unknown. In this study, we aimed to determine which UV wavelengths are effective to expose Gafchromic XR‐RV3 and XR‐SP2. UV lamps with peak wavelengths of 245 nm, 310 nm, and 365 nm were used. The three UV wavelengths were used to irradiate Gafchromic XR‐RV3 and XR‐SP2 films for 60 min, and irradiation was repeated every 60 min for 600 min thereafter. Films were scanned after each irradiation period on a flatbed scanner. The images were split into their red‐green‐blue components, and red images were stored using ImageJ version 1.44o image analysis software. Regions of interest (ROI), 0.5 inches in diameter, were placed at the centers of the subtracted Gafchromic film images, and graphs of UV irradiation duration and mean pixel values were plotted. There were reactions to UV‐A on both Gafchromic XR‐RV3 and XR‐SP2; those to UV‐B were moderate. However, UV‐C demonstrated few reactions with Gafchromic XR‐RV3 and XR‐SP2. From these results, irradiation with UV‐A may be able to correct nonuniformity errors. Uniform UV‐A irradiation of Gafchromic films with large areas is possible, and UV rays can be used as a substitute for X‐rays in the double exposure technique. PACS number: 87.53 Bn PMID:26699326

  18. Application of a color scanner for 60Co high dose rate brachytherapy dosimetry with EBT radiochromic film

    PubMed Central

    Ghorbani, Mahdi; Toossi, Mohammad Taghi Bahreyni; Mowlavi, Ali Asghar; Roodi, Shahram Bayani; Meigooni, Ali Soleimani

    2012-01-01

    Background. The aim of this study is to evaluate the performance of a color scanner as a radiochromic film reader in two dimensional dosimetry around a high dose rate brachytherapy source. Materials and methods A Microtek ScanMaker 1000XL film scanner was utilized for the measurement of dose distribution around a high dose rate GZP6 60Co brachytherapy source with GafChromic® EBT radiochromic films. In these investigations, the non-uniformity of the film and scanner response, combined, as well as the films sensitivity to scanner’s light source was evaluated using multiple samples of films, prior to the source dosimetry. The results of these measurements were compared with the Monte Carlo simulated data using MCNPX code. In addition, isodose curves acquired by radiochromic films and Monte Carlo simulation were compared with those provided by the GZP6 treatment planning system. Results Scanning of samples of uniformly irradiated films demonstrated approximately 2.85% and 4.97% nonuniformity of the response, respectively in the longitudinal and transverse directions of the film. Our findings have also indicated that the film response is not affected by the exposure to the scanner’s light source, particularly in multiple scanning of film. The results of radiochromic film measurements are in good agreement with the Monte Carlo calculations (4%) and the corresponding dose values presented by the GZP6 treatment planning system (5%). Conclusions The results of these investigations indicate that the Microtek ScanMaker 1000XL color scanner in conjunction with GafChromic EBT film is a reliable system for dosimetric evaluation of a high dose rate brachytherapy source. PMID:23411947

  19. Technical Note: Magnetic field effects on Gafchromic-film response in MR-IGRT.

    PubMed

    Reynoso, Francisco J; Curcuru, Austen; Green, Olga; Mutic, Sasa; Das, Indra J; Santanam, Lakshmi

    2016-12-01

    Magnetokinetic changes may affect crystal orientation and polymerization within the active layer of radiochromic film (RCF). This effect is investigated in a magnetic resonance image-guided radiotherapy unit within the context of film dosimetry. Gafchromic EBT2 RCF was irradiated in a 30 × 30 × 30 cm 3 solid water phantom using a Co-60 MRI guided radiotherapy system (B = 0.35 T) under normal operating conditions, and under the exact conditions and setup without a magnetic field. Fifteen 20.3 × 25.4 cm 2 EBT2 film sheets were placed at three different depths (d = 0.5, 5, and 10 cm) using five different treatment plans. The plans were computed using the MRIdian (ViewRay, Inc.) treatment planning system to deliver doses between 0 and 17.6 Gy. Films were analyzed before and after irradiation to obtain the net optical density (netOD) for each color channel separately. Scanning electron microscope (SEM) images were obtained to compare the active layer of selected samples. The results indicated that the red channel netOD decreased between 0.013 and 0.123 (average of 0.060 ± 0.033) for doses above 2.8 Gy, with a linear increase in this effect for higher doses. Green channel netOD showed similar results with a decrease between 0.012 and 0.105 (average of 0.041 ± 0.027) for doses above 3.5 Gy. The blue channel showed the weakest effect with a netOD decrease between 0.013 and 0.029 (average of 0.020 ± 0.006) for doses above 8.0 Gy. SEM images show changes in crystal orientation within active layer in RCF exposed in a magnetic field. The presence of a magnetic field affects crystal orientation and polymerization during irradiation, where netOD decreased by an average of 8.7%, 8.0%, and 4.3% in the red, green, and blue channels, respectively. The under response was dependent on dose and differed by up to 15% at 17.6 Gy.

  20. Investigation of EBT3 radiochromic film's response to humidity.

    PubMed

    León-Marroquín, Elsa Y; Lárraga-Gutiérrez, José M; Herrera-González, J Alfredo; Camacho-López, Miguel A; Villarreal Barajas, José E; García-Garduño, Olivia A

    2018-05-01

    The aim of this work is to investigate the effects of immersing EBT3 radiochromic film in water and to evaluate its contribution to the total uncertainty in dose determination. We used 3 cm × 3 cm EBT3 radiochromic films irradiated in the range of 0-70 Gy to study the impact of water immersion on the change in net optical density. These films were placed in a water container for a period of 24 h. The net optical density was measured before (0 h) and after of the immersion in water (1, 3, 6, 12, 18, and 24 h). The absorbance spectrum of the EBT3 radiochromic film was measured at 0 h and 24 h after immersion in water. The uncertainty in dose determination due to the effects of keeping the EBT3 radiochromic film submerged in water at 0, 1, and 24 h were recorded in the red, green, and blue channels. We observed an increase in the net optical density as an effect on the film due to its immersion in water. The penetration of the water at the edges of the radiochromic film was observed to be a function of time during which the film remained in the water. On the other hand, the penetration of water at the edges of the film was found to be independent of irradiation dose. EBT3 radiochromic film is found more resistant to water penetration through the edges than its predecessors. However, there is evidence that suggest that liquid water damage the Nylon cover layer of the film by changing its optical properties. Therefore, it is recommended to build a new calibration curve for radiochromic films for a specific situation involving dose measurements in liquid water. © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  1. How flatbed scanners upset accurate film dosimetry

    NASA Astrophysics Data System (ADS)

    van Battum, L. J.; Huizenga, H.; Verdaasdonk, R. M.; Heukelom, S.

    2016-01-01

    Film is an excellent dosimeter for verification of dose distributions due to its high spatial resolution. Irradiated film can be digitized with low-cost, transmission, flatbed scanners. However, a disadvantage is their lateral scan effect (LSE): a scanner readout change over its lateral scan axis. Although anisotropic light scattering was presented as the origin of the LSE, this paper presents an alternative cause. Hereto, LSE for two flatbed scanners (Epson 1680 Expression Pro and Epson 10000XL), and Gafchromic film (EBT, EBT2, EBT3) was investigated, focused on three effects: cross talk, optical path length and polarization. Cross talk was examined using triangular sheets of various optical densities. The optical path length effect was studied using absorptive and reflective neutral density filters with well-defined optical characteristics (OD range 0.2-2.0). Linear polarizer sheets were used to investigate light polarization on the CCD signal in absence and presence of (un)irradiated Gafchromic film. Film dose values ranged between 0.2 to 9 Gy, i.e. an optical density range between 0.25 to 1.1. Measurements were performed in the scanner’s transmission mode, with red-green-blue channels. LSE was found to depend on scanner construction and film type. Its magnitude depends on dose: for 9 Gy increasing up to 14% at maximum lateral position. Cross talk was only significant in high contrast regions, up to 2% for very small fields. The optical path length effect introduced by film on the scanner causes 3% for pixels in the extreme lateral position. Light polarization due to film and the scanner’s optical mirror system is the main contributor, different in magnitude for the red, green and blue channel. We concluded that any Gafchromic EBT type film scanned with a flatbed scanner will face these optical effects. Accurate dosimetry requires correction of LSE, therefore, determination of the LSE per color channel and dose delivered to the film.

  2. Outdoor solar UVA dose assessment with EBT2 radiochromic film using spectrophotometer and densitometer measurements.

    PubMed

    Abukassem, I; Bero, M A

    2015-04-01

    Direct measurements of solar ultraviolet radiations (UVRs) have an important role in the protection of humans against UVR hazard. This work presents simple technique based on the application of EBT2 GAFCHROMIC(®) film for direct solar UVA dose assessment. It demonstrates the effects of different parts of the solar spectrum (UVB, visible and infrared) on performed UVA field measurements and presents the measurement uncertainty budget. The gradient of sunlight exposure level permitted the authors to establish the mathematical relationships between the measured solar UVA dose and two measured quantities: the first was the changes in spectral absorbance at the wavelength 633 nm (A633) and the second was the optical density (OD). The established standard relations were also applied to calculate the solar UVA dose variations during the whole day; 15 min of exposure each hour between 8:00 and 17:00 was recorded. Results show that both applied experimental methods, spectrophotometer absorbance and densitometer OD, deliver comparable figures for EBT2 solar UVA dose assessment with relative uncertainty of 11% for spectral absorbance measurements and 15% for OD measurements. © The Author 2014. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  3. Local heterogeneities in early batches of EBT2 film: a suggested solution.

    PubMed

    Kairn, T; Aland, T; Kenny, J

    2010-08-07

    To enhance the utility of radiochromic films for high-resolution dosimetry of small and modulated radiotherapy fields, we propose a means to negate the effects of heterogeneities in EBT2 (and other) films. The results of using our simple procedure for evaluating radiation dose in EBT2 film are compared with the results of using the manufacturer's recommended procedure as well as a procedure previously established for evaluating dose in older EBT film. It is shown that Newton's ring-like scanning artefacts can be avoided through the use of a plastic frame, to elevate the film above the scanner's surface. The effects of film heterogeneity can be minimized by evaluating net optical density, pixelwise, as the logarithm of the ratio of the red-channel pixel value in each pixel of each irradiated film to the red-channel pixel value in the same pixel in the same film prior to irradiation. The application of a blue-channel correction was found to result in increased noise. It is recommended that, when using EBT2 film for radiotherapy quality assurance, the films should be scanned before and after irradiation and analysed using the method proposed herein, without the use of the blue-channel correction, in order to produce dose images with minimal film heterogeneity effects.

  4. Spectral calibration of EBT3 and HD-V2 radiochromic film response at high dose using 20 MeV proton beams

    NASA Astrophysics Data System (ADS)

    Feng, Yiwei; Tiedje, Henry F.; Gagnon, Katherine; Fedosejevs, Robert

    2018-04-01

    Radiochromic film is used extensively in many medical, industrial, and scientific applications. In particular, the film is used in analysis of proton generation and in high intensity laser-plasma experiments where very high dose levels can be obtained. The present study reports calibration of the dose response of Gafchromic EBT3 and HD-V2 radiochromic films up to high exposure densities. A 2D scanning confocal densitometer system is employed to carry out accurate optical density measurements up to optical density 5 on the exposed films at the peak spectral absorption wavelengths. Various wavelengths from 400 to 740 nm are also scanned to extend the practical dose range of such films by measuring the response at wavelengths removed from the peak response wavelengths. Calibration curves for the optical density versus exposure dose are determined and can be used for quantitative evaluation of measured doses based on the measured optical densities. It was found that blue and UV wavelengths allowed the largest dynamic range though at some trade-off with overall accuracy.

  5. Poster — Thur Eve — 53: Novel Technique for the Measurement of Ultra-Superficial Doses Using Gafchromic Film

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Marcos, M.; Devic, S.

    2014-08-15

    Purpose: Dose build-up and electron contamination are two closely related quantities with important implications in radiotherapy, yet they are quite difficult to measure with great certainty. We present a novel technique for measuring ultra-superficial doses. Method and Materials: We used Gafchromic EBT-3 film which have an effective point of measurement of roughly 153 micros (effective depth in water). By peeling off one of the polyester layers, the active layer becomes the top layer and we obtain a film with a effective point of measurement of 15 microns (effective depth in water). A film calibration was performed using a 180 kVpmore » orthovoltage beam. Since the active layer of the film may have been compressed or perturbed during the peeling of clear polyester we use a triple-channel film calibration technique to minimize the effects of non-uniformity in the active layer. We measured surface doses of orthovoltage beams with lead cutouts in place to introduce contaminant photoelectrons. Results: Our measurements show that the dose enhancement near the edges of the lead were about 125% relative to central axis for 6 cm diameter cutouts up to 170% for 2 cm diameter cutouts, which were within 5% of our EGSnrc based Monte Carlo simulations.« less

  6. SU-E-T-135: Investigation of Commercial-Grade Flatbed Scanners and a Medical- Grade Scanner for Radiochromic EBT Film Dosimetry.

    PubMed

    Syh, J; Patel, B; Syh, J; Wu, H; Rosen, L; Durci, M; Katz, S; Sibata, C

    2012-06-01

    To evaluate the characteristics of commercial-grade flatbed scanners and medical-grade scanners for radiochromic EBT film dosimetry. Performance aspects of a Vidar Dosimetry Pro Advantage (Red), Epson 750 Pro, Microtek ArtixScan 1800f, and Microtek ScanMaker 8700 scanner for EBT2 Gafchromic film were evaluated in the categories of repeatability, maximum distinguishable optical density (OD) differentiation, OD variance, and dose curve characteristics. OD step film by Stouffer Industries containing 31 steps ranging from 0.05 to 3.62 OD was used. EBT films were irradiated with dose ranging from 20 to 600 cGy in 6×6 cm 2 field sizes and analyzed 24 hours later using RIT113 and Tomotherapy Film Analyzer software. Scans were performed in transmissive mode, landscape orientation, 16-bit image. The mean and standard deviation Analog to Digital (A/D) scanner value was measured by selecting a 3×3 mm 2 uniform area in the central region of each OD step from a total of 20 scans performed over several weeks. Repeatability was determined from the variance of OD step 0.38. Maximum distinguishable OD was defined as the last OD step whose range of A/D values does not overlap with its neighboring step. Repeatability uncertainty ranged from 0.1% for Vidar to 4% for Epson. Average standard deviation of OD steps ranged from 0.21% for Vidar to 6.4% for ArtixScan 1800f. Maximum distinguishable optical density ranged from 3.38 for Vidar to 1.32 for ScanMaker 8700. A/D range of each OD step corresponds to a dose range. Dose ranges of OD steps varied from 1% for Vidar to 20% for ScanMaker 8700. The Vidar exhibited a dose curve that utilized a broader range of OD values than the other scanners. Vidar exhibited higher maximum distinguishable OD, smaller variance in repeatability, smaller A/D value deviation per OD step, and a shallower dose curve with respect to OD. © 2012 American Association of Physicists in Medicine.

  7. Gamma index evaluation of IMRT technique using gafchromic film EBT3 for homogeneous and inhomogeneous material

    NASA Astrophysics Data System (ADS)

    Nisauf, T. A.; Wibowo, W. E.; Pawiro, S. A.

    2017-07-01

    This study was done to evaluate the gamma index for registering between the planar of dose planning and the measurement of EBT film. The treatment plan was simulated for 5 patients using Fan Beam Computerized Tomography (FBCT) modality, Philips Pinnacle planning system, 6 MV photon energy, 50 segments IMRT technique, and calculation grid resolution (CGR) of 0.2 cm. Gamma Index (GI) evaluation was done with criteria of dose difference (DD) of 2 %, dose to agreement (DTA) of 2 mm and dose difference (DD) of 5 % DTA of 3 mm, SAD 100 cm, depth of 5 cm and 10 cm of the phantom. The result shows that GI for homogeneous material is greater than for inhomogeneous material with discrepancy to previous work is about 1.98 % for homogeneous material (depth 5 cm) and 2.05 % (depth 10 cm) while it was found of 2.98 % for inhomogeneous material (equivalent depth 5 cm) and 4.59 % (equivalent depth 10 cm).

  8. Design and Calibration of a X-Ray Millibeam

    DTIC Science & Technology

    2005-12-01

    developed for use in Fricke dosimetry , parallel-plate ionization chambers, Lithium Fluoride thermoluminescent dosimetry ( TLD ), and EBT GafChromic...thermoluminescent dosimetry ( TLD ), and EBT GafChromic film to characterize the spatial distribution and accuracy of the doses produced by the Faxitron. A...absorbed dose calibration factors for use in Fricke dosimetry , parallel-plate ionization chambers, Lithium Fluoride (LiF) TLD , and EBT GafChromic film. The

  9. Calibration of EBT2 film using a red-channel PDD method in combination with a modified three-channel technique.

    PubMed

    Chang, Liyun; Ho, Sheng-Yow; Lee, Tsair-Fwu; Yeh, Shyh-An; Ding, Hueisch-Jy; Chen, Pang-Yu

    2015-10-01

    Ashland Inc. EBT2 and EBT3 films are widely used in quality assurance for radiation therapy; however, there remains a relatively high degree of uncertainty [B. Hartmann, M. Martisikova, and O. Jakel, "Homogeneity of Gafchromic EBT2 film," Med. Phys. 37, 1753-1756 (2010)]. Micke et al. (2011) recently improved the spatial homogeneity using all color channels of a flatbed scanner; however, van Hoof et al. (2012) pointed out that the corrected nonuniformity still requires further investigation for larger fields. To reduce the calibration errors and the uncertainty, the authors propose a new red-channel percentage-depth-dose method in combination with a modified three-channel technique. For the ease of comparison, the EBT2 film image used in the authors' previous study (2012) was reanalyzed using different approaches. Photon beams of 6-MV were delivered to two different films at two different beam on times, resulting in the absorption doses of ranging from approximately 30 to 300 cGy at the vertical midline of the film, which was set to be coincident with the central axis of the beam. The film was tightly sandwiched in a 30(3)-cm(3) polystyrene phantom, and the pixel values for red, green, and blue channels were extracted from 234 points on the central axis of the beam and compared with the corresponding depth doses. The film was first calibrated using the multichannel method proposed by Micke et al. (2010), accounting for nonuniformities in the scanner. After eliminating the scanner and dose-independent nonuniformities, the film was recalibrated via the dose-dependent optical density of the red channel and fitted to a power function. This calibration was verified via comparisons of the dose profiles extracted from the films, where three were exposed to a 60° physical wedge field and three were exposed to composite fields, and all of which were measured in a water phantom. A correction for optical attenuation was implemented, and treatment plans of intensity modulated

  10. Neutron response of GafChromic® EBT2 film

    NASA Astrophysics Data System (ADS)

    Hsiao, Ming-Chen; Liu, Yuan-Hao; Chen, Wei-Lin; Jiang, Shiang-Huei

    2013-03-01

    Neutron and gamma-ray mixed field dosimetry remains one of the most challenging topics in radiation dosimetry studies. However, the requirement for accurate mixed field dosimetry is increasing because of the considerable interest in high-energy radiotherapy machines, medical ion beams and BNCT epithermal neutron beams. Therefore, this study investigated the GafChromic® EBT2 film. The linearity, reproducibility, energy dependence and homogeneity of the film were tested in a 60Co medical beam, 6-MV LINAC and 10-MV LINAC. The linearity and self-developing effect of the film irradiated in an epithermal neutron beam were also examined. These basic detector characteristics showed that EBT2 film can be effectively applied in mixed field dosimetry. A general detector response model was developed to determine the neutron relative effectiveness (RE) values. The RE value of fast neutrons varies with neutron spectra. By contrast, the RE value of thermal neutrons was determined as a constant; it is only 32.5% in relation to gamma rays. No synergy effect was observed in this study. The lithium-6 capture reaction dominates the neutron response in the thermal neutron energy range, and the recoil hydrogen dose becomes the dominant component in the fast neutron energy region. Based on this study, the application of the EBT2 film in the neutron and gamma-ray mixed field is feasible.

  11. Development of the optimal radiochromic film dosimetry system for measurement of IMRT radiation beams

    NASA Astrophysics Data System (ADS)

    Baker, Jameson Todd

    The complex dose patterns that result in Intensity Modulated Radiation Therapy make the typical QA of a second calculation insufficient for ensuring safe treatment of patients. Many facilities choose to deliver the treatment to film inserted in a phantom and calculate the dose delivered as an additional check of the treatment plan. Radiochromic films allow for measurements without the use of a processor in the current digital age. International Specialty Products developed Gafchromic EBT film, which is a radiochromic film having a useful range of 1 -- 800 cGy. EBT film properties are fully analyzed including studies of uniformity, spectral absorption, exposure sensitivity, energy dependence and post exposure density growth. Dosimetric performance on commercially available digitizers is studied with specific attention on the shortcomings. Finally, a custom designed scanner is built specifically for EBT film and its unique properties. Performance of the EBT digitizer is analyzed and compared against currently available scanners.

  12. A spectroscopic study of the chromatic properties of GafChromicEBT3 films.

    PubMed

    Callens, M; Crijns, W; Simons, V; De Wolf, I; Depuydt, T; Maes, F; Haustermans, K; D'hooge, J; D'Agostino, E; Wevers, M; Pfeiffer, H; Van Den Abeele, K

    2016-03-01

    This work provides an interpretation of the chromatic properties of GafChromicEBT3 films based on the chemical nature of the polydiacetylene (PDA) molecules formed upon interaction with ionizing radiation. The EBT3 films become optically less transparent with increasing radiation dose as a result of the radiation-induced polymerization of diacetylene monomers. In contrast to empirical quantification of the chromatic properties, less attention has been given to the underlying molecular mechanism that induces the strong decrease in transparency. Unlaminated GafChromicEBT3 films were irradiated with a 6 MV photon beam to dose levels up to 20 Gy. The optical absorption properties of the films were investigated using visible (vis) spectroscopy. The presence of PDA molecules in the active layer of the EBT3 films was investigated using Raman spectroscopy, which probes the vibrational modes of the molecules in the layer. The vibrational modes assigned to PDA's were used in a theoretical vis-absorption model to fit our experimental vis-absorption spectra. From the fit parameters, one can assess the relative contribution of different PDA conformations and the length distribution of PDA's in the film. Vis-spectroscopy shows that the optical density increases with dose in the full region of the visible spectrum. The Raman spectrum is dominated by two vibrational modes, most notably by the ν(C≡C) and the ν(C=C) stretching modes of the PDA backbone. By fitting the vis-absorption model to experimental spectra, it is found that the active layer contains two distinct PDA conformations with different absorption properties and reaction kinetics. Furthermore, the mean PDA conjugation length is found to be 2-3 orders of magnitude smaller than the crystals PDA's are embedded in. Vis- and Raman spectroscopy provided more insight into the molecular nature of the radiochromic properties of EBT3 films through the identification of the excited states of PDA and the presence of two PDA

  13. Calibration of EBT2 film using a red-channel PDD method in combination with a modified three-channel technique

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chang, Liyun, E-mail: liyunc@isu.edu.tw; Ho, Sheng-Yow; Lee, Tsair-Fwu

    Purpose: Ashland Inc. EBT2 and EBT3 films are widely used in quality assurance for radiation therapy; however, there remains a relatively high degree of uncertainty [B. Hartmann, M. Martisikova, and O. Jakel, “Homogeneity of Gafchromic EBT2 film,” Med. Phys. 37, 1753–1756 (2010)]. Micke et al. (2011) recently improved the spatial homogeneity using all color channels of a flatbed scanner; however, van Hoof et al. (2012) pointed out that the corrected nonuniformity still requires further investigation for larger fields. To reduce the calibration errors and the uncertainty, the authors propose a new red-channel percentage-depth-dose method in combination with a modified three-channelmore » technique. Methods: For the ease of comparison, the EBT2 film image used in the authors’ previous study (2012) was reanalyzed using different approaches. Photon beams of 6-MV were delivered to two different films at two different beam on times, resulting in the absorption doses of ranging from approximately 30 to 300 cGy at the vertical midline of the film, which was set to be coincident with the central axis of the beam. The film was tightly sandwiched in a 30{sup 3}-cm{sup 3} polystyrene phantom, and the pixel values for red, green, and blue channels were extracted from 234 points on the central axis of the beam and compared with the corresponding depth doses. The film was first calibrated using the multichannel method proposed by Micke et al. (2010), accounting for nonuniformities in the scanner. After eliminating the scanner and dose-independent nonuniformities, the film was recalibrated via the dose-dependent optical density of the red channel and fitted to a power function. This calibration was verified via comparisons of the dose profiles extracted from the films, where three were exposed to a 60° physical wedge field and three were exposed to composite fields, and all of which were measured in a water phantom. A correction for optical attenuation was implemented

  14. Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film.

    PubMed

    Sonier, Marcus; Wronski, Matt; Yeboah, Collins

    2015-03-08

    Lens dose is a concern during the treatment of facial lesions with anterior electron beams. Lead shielding is routinely employed to reduce lens dose and minimize late complications. The purpose of this work is twofold: 1) to measure dose pro-files under large-area lead shielding at the lens depth for clinical electron energies via film dosimetry; and 2) to assess the accuracy of the Pinnacle treatment planning system in calculating doses under lead shields. First, to simulate the clinical geometry, EBT3 film and 4 cm wide lead shields were incorporated into a Solid Water phantom. With the lead shield inside the phantom, the film was positioned at a depth of 0.7 cm below the lead, while a variable thickness of solid water, simulating bolus, was placed on top. This geometry was reproduced in Pinnacle to calculate dose profiles using the pencil beam electron algorithm. The measured and calculated dose profiles were normalized to the central-axis dose maximum in a homogeneous phantom with no lead shielding. The resulting measured profiles, functions of bolus thickness and incident electron energy, can be used to estimate the lens dose under various clinical scenarios. These profiles showed a minimum lead margin of 0.5 cm beyond the lens boundary is required to shield the lens to ≤ 10% of the dose maximum. Comparisons with Pinnacle showed a consistent overestimation of dose under the lead shield with discrepancies of ~ 25% occur-ring near the shield edge. This discrepancy was found to increase with electron energy and bolus thickness and decrease with distance from the lead edge. Thus, the Pinnacle electron algorithm is not recommended for estimating lens dose in this situation. The film measurements, however, allow for a reasonable estimate of lens dose from electron beams and for clinicians to assess the lead margin required to reduce the lens dose to an acceptable level.

  15. Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film

    PubMed Central

    Wronski, Matt; Yeboah, Collins

    2015-01-01

    Lens dose is a concern during the treatment of facial lesions with anterior electron beams. Lead shielding is routinely employed to reduce lens dose and minimize late complications. The purpose of this work is twofold: 1) to measure dose profiles under large‐area lead shielding at the lens depth for clinical electron energies via film dosimetry; and 2) to assess the accuracy of the Pinnacle treatment planning system in calculating doses under lead shields. First, to simulate the clinical geometry, EBT3 film and 4 cm wide lead shields were incorporated into a Solid Water phantom. With the lead shield inside the phantom, the film was positioned at a depth of 0.7 cm below the lead, while a variable thickness of solid water, simulating bolus, was placed on top. This geometry was reproduced in Pinnacle to calculate dose profiles using the pencil beam electron algorithm. The measured and calculated dose profiles were normalized to the central‐axis dose maximum in a homogeneous phantom with no lead shielding. The resulting measured profiles, functions of bolus thickness and incident electron energy, can be used to estimate the lens dose under various clinical scenarios. These profiles showed a minimum lead margin of 0.5 cm beyond the lens boundary is required to shield the lens to ≤10% of the dose maximum. Comparisons with Pinnacle showed a consistent overestimation of dose under the lead shield with discrepancies of ∼25% occurring near the shield edge. This discrepancy was found to increase with electron energy and bolus thickness and decrease with distance from the lead edge. Thus, the Pinnacle electron algorithm is not recommended for estimating lens dose in this situation. The film measurements, however, allow for a reasonable estimate of lens dose from electron beams and for clinicians to assess the lead margin required to reduce the lens dose to an acceptable level. PACS number(s): 87.53.Bn, 87.53.Kn, 87.55.‐x, 87.55.D‐ PMID:27074448

  16. Reference dosimetry using radiochromic film

    PubMed Central

    Girard, Frédéric; Bouchard, Hugo

    2012-01-01

    The objectives of this study are to identify and quantify factors that influence radiochromic film dose response and to determine whether such films are suitable for reference dosimetry. The influence of several parameters that may introduce systematic dose errors when performing reference dose measurements were investigated. The effect of the film storage temperature was determined by comparing the performance of three lots of GAFCHROMIC EBT2 films stored at either 4°C or room temperature. The effect of high (>80%) or low (<20%) relative humidity was also determined. Doses measured in optimal conditions with EBT and EBT2 films were then compared with an A12 ionization chamber measurement. Intensity‐modulated radiation therapy quality controls using EBT2 films were also performed in reference dose. The results obtained using reference dose measurements were compared with those obtained using relative dose measurements. Storing the film at 4°C improves the stability of the film over time, but does not eliminate the noncatalytic film development, seen as a rise in optical density over time in the absence of radiation. Relative humidity variations ranging from 80% to 20% have a strong impact on the optical density and could introduce dose errors of up to 15% if the humidity were not controlled during the film storage period. During the scanning procedure, the film temperature influences the optical density that is measured. When controlling for these three parameters, the dose differences between EBT or EBT2 and the A12 chamber are found to be within ±4% (2σ level) over a dose range of 20–350 cGy. Our results also demonstrate the limitation of the Anisotropic Analytical Algorithm for dose calculation of highly modulated treatment plans. PACS numbers: 87.55.Qr; 87.56.Fc PMID:23149793

  17. Dependency of EBT2 film calibration curve on postirradiation time

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chang, Liyun, E-mail: liyunc@isu.edu.tw; Ding, Hueisch-Jy; Ho, Sheng-Yow

    2014-02-15

    Purpose: The Ashland Inc. product EBT2 film model is a widely used quality assurance tool, especially for verification of 2-dimensional dose distributions. In general, the calibration film and the dose measurement film are irradiated, scanned, and calibrated at the same postirradiation time (PIT), 1-2 days after the films are irradiated. However, for a busy clinic or in some special situations, the PIT for the dose measurement film may be different from that of the calibration film. In this case, the measured dose will be incorrect. This paper proposed a film calibration method that includes the effect of PIT. Methods: Themore » dose versus film optical density was fitted to a power function with three parameters. One of these parameters was PIT dependent, while the other two were found to be almost constant with a standard deviation of the mean less than 4%. The PIT-dependent parameter was fitted to another power function of PIT. The EBT2 film model was calibrated using the PDD method with 14 different PITs ranging from 1 h to 2 months. Ten of the fourteen PITs were used for finding the fitting parameters, and the other four were used for testing the model. Results: The verification test shows that the differences between the delivered doses and the film doses calculated with this modeling were mainly within 2% for delivered doses above 60 cGy, and the total uncertainties were generally under 5%. The errors and total uncertainties of film dose calculation were independent of the PIT using the proposed calibration procedure. However, the fitting uncertainty increased with decreasing dose or PIT, but stayed below 1.3% for this study. Conclusions: The EBT2 film dose can be modeled as a function of PIT. For the ease of routine calibration, five PITs were suggested to be used. It is recommended that two PITs be located in the fast developing period (1∼6 h), one in 1 ∼ 2 days, one around a week, and one around a month.« less

  18. NOTE: Dose area product evaluations with Gafchromic® XR-R films and a flat-bed scanner

    NASA Astrophysics Data System (ADS)

    Rampado, O.; Garelli, E.; Deagostini, S.; Ropolo, R.

    2006-12-01

    Gafchromic® XR-R films are a useful tool to evaluate entrance skin dose in interventional radiology. Another dosimetric quantity of interest in diagnostic and interventional radiology is the dose area product (DAP). In this study, a method to evaluate DAP using Gafchromic® XR-R films and a flat-bed scanner was developed and tested. Film samples were exposed to an x-ray beam of 80 kVp over a dose range of 0 10 Gy. DAP measurements with films were obtained from the digitalization of a film sample positioned over the x-ray beam window during the exposure. DAP values obtained with this method were compared for 23 cardiological interventional procedures with DAP values displayed by the equipment. The overall one-sigma dose measurement uncertainty depended on the absorbed dose, with values below 6% for doses above 1 Gy. A maximum discrepancy of 16% was found, which is of the order of the differences in the DAP measurements that may occur with different calibration procedures. Based on the results presented, after an accurate calibration procedure and a thorough inspection of the relationship between the actual dose and the direct measured quantity (net optical density or net pixel value variation), Gafchromic® XR-R films can be used to assess the DAP.

  19. Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density.

    PubMed

    Marroquin, Elsa Y León; Herrera González, José A; Camacho López, Miguel A; Barajas, José E Villarreal; García-Garduño, Olivia A

    2016-09-08

    Radiochromic film has become an important tool to verify dose distributions for intensity-modulated radiotherapy (IMRT) and quality assurance (QA) procedures. A new radiochromic film model, EBT3, has recently become available, whose composition and thickness of the sensitive layer are the same as those of previous EBT2 films. However, a matte polyester layer was added to EBT3 to prevent the formation of Newton's rings. Furthermore, the symmetrical design of EBT3 allows the user to eliminate side-orientation dependence. This film and the flatbed scanner, Epson Perfection V750, form a dosimetry system whose intrinsic characteristics were studied in this work. In addition, uncertainties associated with these intrinsic characteristics and the total uncertainty of the dosimetry system were determined. The analysis of the response of the radiochromic film (net optical density) and the fitting of the experimental data to a potential function yielded an uncertainty of 2.6%, 4.3%, and 4.1% for the red, green, and blue channels, respectively. In this work, the dosimetry system presents an uncertainty in resolving the dose of 1.8% for doses greater than 0.8 Gy and less than 6 Gy for red channel. The films irradiated between 0 and 120 Gy show differences in the response when scanned in portrait or landscape mode; less uncertainty was found when using the portrait mode. The response of the film depended on the position on the bed of the scanner, contributing an uncertainty of 2% for the red, 3% for the green, and 4.5% for the blue when placing the film around the center of the bed of scanner. Furthermore, the uniformity and reproducibility radiochromic film and reproducibility of the response of the scanner contribute less than 1% to the overall uncertainty in dose. Finally, the total dose uncertainty was 3.2%, 4.9%, and 5.2% for red, green, and blue channels, respectively. The above uncertainty values were obtained by mini-mizing the contribution to the total dose uncertainty

  20. Characterization of a Gafchromic film for the two-dimensional profile measurement of low-energy heavy-ion beams

    NASA Astrophysics Data System (ADS)

    Yuri, Yosuke; Narumi, Kazumasa; Yuyama, Takahiro

    2016-08-01

    The feasibility of the transverse intensity distribution measurement of low-energy (keV/u range) heavy-ion beams using radiochromic films is experimentally explored. We employ a Gafchromic radiochromic film, HD-V2, whose active layer is not laminated by a surface-protection layer. The coloration response of films irradiated with several ion beams is characterized in terms of optical density (OD) by reading the films with a general-purpose scanner. To explore the energy dependence of the film response widely, the kinetic energy of the beams is varied from 1.5 keV/u to 27 MeV/u. We have found that the coloration of HD-V2 films is induced by irradiation with low-energy ion beams of the order of 10 keV/u. The range of the beams is considerably shorter than the thickness of the film's active layer. The dependence of OD response on ion species is also discussed. We demonstrate that the Gafchromic film used here is useful for measuring the intensity distribution of such low-energy ion beams.

  1. Magnetic fields are causing small, but significant changes of the radiochromic EBT3 film response to 6 MV photons.

    PubMed

    Delfs, Björn; Schoenfeld, Andreas A; Poppinga, Daniela; Kapsch, Ralf-Peter; Jiang, Ping; Harder, Dietrich; Poppe, Björn; Looe, Hui Khee

    2018-01-31

    The optical density (OD) of EBT3 radiochromic films (Ashland Specialty Ingredients, Bridgewater, NJ, USA) exposed to absorbed doses to water up to D  =  20 Gy in magnetic fields of B  =  0.35 and 1.42 T was measured in the three colour channels of an Epson Expression 10000XL flatbed scanner. A 7 cm wide water phantom with fixed film holder was placed between the pole shoes of a constant-current electromagnet with variable field strength and was irradiated by a 6 MV photon beam whose axis was directed at right angles with the field lines. The doses at the film position at water depth 5 cm were measured with a calibrated ionization chamber when the magnet was switched off and were converted to the doses in presence of the magnetic field via the monitor units and by a Monte Carlo-calculated correction accounting for the slight change of the depth dose curves in magnetic fields. In the presence of the 0.35 and 1.42 T fields small negative changes of the OD values at given absorbed doses to water occurred and just significantly exceeded the uncertainty margin given by the stochastic and the uncorrected systematic deviations. This change can be described by a  +2.1% change of the dose values needed to produce a given optical density in the presence of a 1.42 T field. The thereby modified OD versus D function remained unchanged irrespective of whether the original short film side-the preference direction of the monomer crystals of the film-was directed parallel or orthogonal to the magnetic field. The 'orientation effect', the difference between the optical densities measured in the 'portrait' or 'landscape' film positions on the scanner bed caused by the reflection of polarised light in the scanner's mirror system, remained unaltered after EBT3 film exposure in magnetic fields. An independent optical bench investigation of EBT3 films exposed to doses of 10 and 20 Gy at 0.35 and 1.42 T showed that the direction of the electric vector of polarised

  2. Poster — Thur Eve — 26: Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sanier, M; Wronski, M; Yeboah, C

    The purpose of this work is twofold: 1) to measure dose profiles under lead shielding at the level of the lens for a range of clinical electron energies via film dosimetry; and, 2) to assess the validity of the Pinnacle treatment planning system (TPS) in calculating the penumbral doses under lead shielding with the heterogeneous electron algorithm. First, a film calibration curve that spanned the electron energies of interest, 6–18MeV, was created. Next, EBT3 film and lead shielding were incorporated into a solid water phantom with the film positioned 7mm below the lead and a variable thickness of bolus onmore » top. This geometry was reproduced in the Pinnacle TPS and used to calculate dose profiles using the heterogeneous electron algorithm. The measured vs. calculated dose profiles were normalized to d{sub max} in a homogeneous phantom with no lead shielding and compared. Pinnacle consistently overestimated the dose distal to the lead shielding with significant discrepancies occurring near the edge of the lead shield reaching 25% at the edge and 35% in the open field region. The film measurements showed that a minimum lead margin of 5mm extending beyond the diameter of the lens is required to adequately shield the lens to ≤10% of the dose at d{sub max}. These measurements allow for a reasonable estimate of the dose to the lens from anterior electron beams. They also allow for clinicians to assess the extent of the lead margin required to reduce the lens dose to an acceptable amount prior to radiotherapy treatment.« less

  3. Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density

    PubMed Central

    Marroquin, Elsa Y. León; Herrera González, José A.; Camacho López, Miguel A.; Barajas, José E. Villarreal

    2016-01-01

    Radiochromic film has become an important tool to verify dose distributions for intensity‐modulated radiotherapy (IMRT) and quality assurance (QA) procedures. A new radiochromic film model, EBT3, has recently become available, whose composition and thickness of the sensitive layer are the same as those of previous EBT2 films. However, a matte polyester layer was added to EBT3 to prevent the formation of Newton's rings. Furthermore, the symmetrical design of EBT3 allows the user to eliminate side‐orientation dependence. This film and the flatbed scanner, Epson Perfection V750, form a dosimetry system whose intrinsic characteristics were studied in this work. In addition, uncertainties associated with these intrinsic characteristics and the total uncertainty of the dosimetry system were determined. The analysis of the response of the radiochromic film (net optical density) and the fitting of the experimental data to a potential function yielded an uncertainty of 2.6%, 4.3%, and 4.1% for the red, green, and blue channels, respectively. In this work, the dosimetry system presents an uncertainty in resolving the dose of 1.8% for doses greater than 0.8 Gy and less than 6 Gy for red channel. The films irradiated between 0 and 120 Gy show differences in the response when scanned in portrait or landscape mode; less uncertainty was found when using the portrait mode. The response of the film depended on the position on the bed of the scanner, contributing an uncertainty of 2% for the red, 3% for the green, and 4.5% for the blue when placing the film around the center of the bed of scanner. Furthermore, the uniformity and reproducibility radiochromic film and reproducibility of the response of the scanner contribute less than 1% to the overall uncertainty in dose. Finally, the total dose uncertainty was 3.2%, 4.9%, and 5.2% for red, green, and blue channels, respectively. The above uncertainty values were obtained by minimizing the contribution to the total dose

  4. Analysis of openings and wide of leaf on multileaf Colimators Using Gafchromic RTQA2 Film

    NASA Astrophysics Data System (ADS)

    Setiawati, Evi; Lailla Rachma, Assyifa; Hidayatullah, M.

    2018-05-01

    The research determined an excitence of correction openings leaf for treatment, and the distribution dose using Gafchromic RTQA2 film. This was about MLC’s correction based on result of movement leaf and field irradiating uniform was done. Methods of research was conduct an irradiating on Gafchromic RTQA2 film based on the index planning homogeneity philosophy, openings leaf and wide leaf. The result of film was lit later in scan. It was continued to include image of the software scanning into matlab. From this case, the image of films common to greyscale image and analysis on the rise in doses blackish films. In this step, we made a correlation between the doses and determine the homogenity to know film dosimetri used homogeneous, and correction of openings leaf and wide leaf. The result between pixel and doses was linear with the equation y = (-0,6)x+108 to low dose and y = (-0,28)x + 108 to high doses and the index of homogeneity range of 0,003 – 0,084. The result homogeneous and correction distribution doses at the openings leaf and wide leaf was around 5% with a value still into the suggested tolerance from ICRU No.50 was 10%.

  5. Magnetic fields are causing small, but significant changes of the radiochromic EBT3 film response to 6 MV photons

    NASA Astrophysics Data System (ADS)

    Delfs, Björn; Schoenfeld, Andreas A.; Poppinga, Daniela; Kapsch, Ralf-Peter; Jiang, Ping; Harder, Dietrich; Poppe, Björn; Khee Looe, Hui

    2018-02-01

    The optical density (OD) of EBT3 radiochromic films (Ashland Specialty Ingredients, Bridgewater, NJ, USA) exposed to absorbed doses to water up to D  =  20 Gy in magnetic fields of B  =  0.35 and 1.42 T was measured in the three colour channels of an Epson Expression 10000XL flatbed scanner. A 7 cm wide water phantom with fixed film holder was placed between the pole shoes of a constant-current electromagnet with variable field strength and was irradiated by a 6 MV photon beam whose axis was directed at right angles with the field lines. The doses at the film position at water depth 5 cm were measured with a calibrated ionization chamber when the magnet was switched off and were converted to the doses in presence of the magnetic field via the monitor units and by a Monte Carlo-calculated correction accounting for the slight change of the depth dose curves in magnetic fields. In the presence of the 0.35 and 1.42 T fields small negative changes of the OD values at given absorbed doses to water occurred and just significantly exceeded the uncertainty margin given by the stochastic and the uncorrected systematic deviations. This change can be described by a  +2.1% change of the dose values needed to produce a given optical density in the presence of a 1.42 T field. The thereby modified OD versus D function remained unchanged irrespective of whether the original short film side—the preference direction of the monomer crystals of the film—was directed parallel or orthogonal to the magnetic field. The ‘orientation effect’, the difference between the optical densities measured in the ‘portrait’ or ‘landscape’ film positions on the scanner bed caused by the reflection of polarised light in the scanner’s mirror system, remained unaltered after EBT3 film exposure in magnetic fields. An independent optical bench investigation of EBT3 films exposed to doses of 10 and 20 Gy at 0.35 and 1.42 T showed that the direction of the electric

  6. Evaluation of a commercial flatbed document scanner and radiographic film scanner for radiochromic EBT film dosimetry

    PubMed Central

    Parker, Brent C.; Neck, Daniel W.; Henkelmann, Greg; Rosen, Isaac I.

    2010-01-01

    The purpose of this study was to quantify the performance and assess the utility of two different types of scanners for radiochromic EBT film dosimetry: a commercial flatbed document scanner and a widely used radiographic film scanner. We evaluated the Epson Perfection V700 Photo flatbed scanner and the Vidar VXR Dosimetry Pro Advantage scanner as measurement devices for radiochromic EBT film. Measurements were made of scan orientation effects, response uniformity, and scanner noise. Scanners were tested using films irradiated with eight separate 3×3 cm2 fields to doses ranging from 0.115–5.119 Gy. ImageJ and RIT software was used for analyzing the Epson and Vidar scans, respectively. For repeated scans of a single film, the measurements in each dose region were reproducible to within ±0.3% standard deviation (SD) with both scanners. Film‐to‐film variations for corresponding doses were measured to be within ±0.4% SD for both Epson scanner and Vidar scanners. Overall, the Epson scanner showed a 10% smaller range of pixel value compared to the Vidar scanner. Scanner noise was small: ±0.3% SD for the Epson and ±0.2% for the Vidar. Overall measurement uniformity for blank film in both systems was better than ±0.2%, provided that the leading and trailing 2 cm film edges were neglected in the Vidar system. In this region artifacts are attributed to the film rollers. Neither system demonstrated a clear measurement advantage. The Epson scanner is a relatively inexpensive method for analyzing radiochromic film, but there is a lack of commercially available software. For a clinic already using a Vidar scanner, applying it to radiochromic film is attractive because commercial software is available. However, care must be taken to avoid using the leading and trailing film edges. PACS number: 87.55.Qr

  7. Use of a Gafchromic film HD-V2 for the profile measurement of energetic ion beams

    NASA Astrophysics Data System (ADS)

    Yuri, Yosuke; Ishizaka, Tomohisa; Agematsu, Takashi; Yuyama, Takahiro; Seito, Hajime; Okumura, Susumu

    2017-09-01

    The coloration response of a radiochromic film, Gafchromic HD-V2, to ion beams was investigated to apply the film to measuring the transverse intensity distribution of large-area ion beams. HD-V2 films were, therefore, irradiated with proton (10 MeV) and several heavy-ion (4.1-27 MeV/u) beams in a wide fluence range at the azimuthally-varying-field cyclotron facility in National Institutes for Quantum and Radiological Science and Technology, and read with an image scanner to analyze changes in the optical density. It was shown that the available fluence range (106-1011 ions/cm2) of HD-V2 depends strongly on ion species, i.e., linear energy transfer (LET). In addition, the reduction of the sensitivity to dose was shown over a wide LET range. The transverse intensity distribution of a large-area ion beam was measured using a response function determined from the measured data. We have demonstrated that the Gafchromic film HD-V2 is useful for measuring the intensity distribution at a low fluence and thus evaluating the characteristics of various ion beams.

  8. Photon beam dosimetry with EBT3 film in heterogeneous regions: Application to the evaluation of dose-calculation algorithms

    NASA Astrophysics Data System (ADS)

    Jung, Hyunuk; Kum, Oyeon; Han, Youngyih; Park, Byungdo; Cheong, Kwang-Ho

    2014-12-01

    For a better understanding of the accuracy of state-of-the-art-radiation therapies, 2-dimensional dosimetry in a patient-like environment will be helpful. Therefore, the dosimetry of EBT3 films in non-water-equivalent tissues was investigated, and the accuracy of commercially-used dose-calculation algorithms was evaluated with EBT3 measurement. Dose distributions were measured with EBT3 films for an in-house-designed phantom that contained a lung or a bone substitute, i.e., an air cavity (3 × 3 × 3 cm3) or teflon (2 × 2 × 2 cm3 or 3 × 3 × 3 cm3), respectively. The phantom was irradiated with 6-MV X-rays with field sizes of 2 × 2, 3 × 3, and 5 × 5 cm2. The accuracy of EBT3 dosimetry was evaluated by comparing the measured dose with the dose obtained from Monte Carlo (MC) simulations. A dose-to-bone-equivalent material was obtained by multiplying the EBT3 measurements by the stopping power ratio (SPR). The EBT3 measurements were then compared with the predictions from four algorithms: Monte Carlo (MC) in iPlan, acuros XB (AXB), analytical anisotropic algorithm (AAA) in Eclipse, and superposition-convolution (SC) in Pinnacle. For the air cavity, the EBT3 measurements agreed with the MC calculation to within 2% on average. For teflon, the EBT3 measurements differed by 9.297% (±0.9229%) on average from the Monte Carlo calculation before dose conversion, and by 0.717% (±0.6546%) after applying the SPR. The doses calculated by using the MC, AXB, AAA, and SC algorithms for the air cavity differed from the EBT3 measurements on average by 2.174, 2.863, 18.01, and 8.391%, respectively; for teflon, the average differences were 3.447, 4.113, 7.589, and 5.102%. The EBT3 measurements corrected with the SPR agreed with 2% on average both within and beyond the heterogeneities with MC results, thereby indicating that EBT3 dosimetry can be used in heterogeneous media. The MC and the AXB dose calculation algorithms exhibited clinically-acceptable accuracy (<5%) in

  9. Dose rate dependence for different dosimeters and detectors: TLD, OSL, EBT films, and diamond detectors.

    PubMed

    Karsch, L; Beyreuther, E; Burris-Mog, T; Kraft, S; Richter, C; Zeil, K; Pawelke, J

    2012-05-01

    The use of laser accelerators in radiation therapy can perhaps increase the low number of proton and ion therapy facilities in some years due to the low investment costs and small size. The laser-based acceleration technology leads to a very high peak dose rate of about 10(11) Gy∕s. A first dosimetric task is the evaluation of dose rate dependence of clinical dosimeters and other detectors. The measurements were done at ELBE, a superconductive linear electron accelerator which generates electron pulses with 5 ps length at 20 MeV. The different dose rates are reached by adjusting the number of electrons in one beam pulse. Three clinical dosimeters (TLD, OSL, and EBT radiochromic films) were irradiated with four different dose rates and nearly the same dose. A faraday cup, an integrating current transformer, and an ionization chamber were used to control the particle flux on the dosimeters. Furthermore two diamond detectors were tested. The dosimeters are dose rate independent up to 4●10(9) Gy∕s within 2% (OSL and TLD) and up to 15●10(9) Gy∕s within 5% (EBT films). The diamond detectors show strong dose rate dependence. TLD, OSL dosimeters, and EBT films are suitable for pulsed beams with a very high pulse dose rate like laser accelerated particle beams.

  10. Dose rate dependence for different dosimeters and detectors: TLD, OSL, EBT films, and diamond detectors

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Karsch, L.; Beyreuther, E.; Burris-Mog, T.

    Purpose: The use of laser accelerators in radiation therapy can perhaps increase the low number of proton and ion therapy facilities in some years due to the low investment costs and small size. The laser-based acceleration technology leads to a very high peak dose rate of about 10{sup 11} Gy/s. A first dosimetric task is the evaluation of dose rate dependence of clinical dosimeters and other detectors. Methods: The measurements were done at ELBE, a superconductive linear electron accelerator which generates electron pulses with 5 ps length at 20 MeV. The different dose rates are reached by adjusting the numbermore » of electrons in one beam pulse. Three clinical dosimeters (TLD, OSL, and EBT radiochromic films) were irradiated with four different dose rates and nearly the same dose. A faraday cup, an integrating current transformer, and an ionization chamber were used to control the particle flux on the dosimeters. Furthermore two diamond detectors were tested. Results: The dosimeters are dose rate independent up to 410{sup 9} Gy/s within 2% (OSL and TLD) and up to 1510{sup 9} Gy/s within 5% (EBT films). The diamond detectors show strong dose rate dependence. Conclusions: TLD, OSL dosimeters, and EBT films are suitable for pulsed beams with a very high pulse dose rate like laser accelerated particle beams.« less

  11. SU-F-T-569: Implementation of a Patient Specific QA Method Using EBT-XD for CyberKnife SRS/SBRT Plans

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zerouali, K; Aubry, J; Doucet, R

    2016-06-15

    Purpose: To implement the new EBT-XD Gafchromic films for accurate dosimetric and geometric validation of stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) CyberKnife (CK) patient specific QA. Methods: Film calibration was performed using a triplechannel film analysis on an Epson 10000XL scanner. Calibration films were irradiated using a Varian Clinac 21EX flattened beam (0 to 20 Gy), to ensure sufficient dose homogeneity. Films were scanned to a resolution of 0.3 mm, 24 hours post irradiation following a well-defined protocol. A set of 12 QA was performed for several types of CK plans: trigeminal neuralgia, brain metastasis, prostate andmore » lung tumors. A custom made insert for the CK head phantom has been manufactured to yield an accurate measured to calculated dose registration. When the high dose region was large enough, absolute dose was also measured with an ionization chamber. Dose calculation is performed using MultiPlan Ray-tracing algorithm for all cases since the phantom is mostly made from near water-equivalent plastic. Results: Good agreement (<2%) was found between the dose to the chamber and the film, when a chamber measurement was possible The average dose difference and standard deviations between film measurements and TPS calculations were respectively 1.75% and 3%. The geometric accuracy has been estimated to be <1 mm, combining robot positioning uncertainty and film registration to calculated dose. Conclusion: Patient specific QA measurements using EBT-XD films yielded a full 2D dose plane with high spatial resolution and acceptable dose accuracy. This method is particularly promising for trigeminal neuralgia plan QA, where the positioning of the spatial dose distribution is equally or more important than the absolute delivered dose to achieve clinical goals.« less

  12. Characterization of differences in calculated and actual measured skin doses to canine limbs during stereotactic radiosurgery using Gafchromic film

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Walters, Jerri; Colorado State University, Fort Collins, CO; Ryan, Stewart

    Accurate calculation of absorbed dose to the skin, especially the superficial and radiosensitive basal cell layer, is difficult for many reasons including, but not limited to, the build-up effect of megavoltage photons, tangential beam effects, mixed energy scatter from support devices, and dose interpolation caused by a finite resolution calculation matrix. Stereotactic body radiotherapy (SBRT) has been developed as an alternative limb salvage treatment option at Colorado State University Veterinary Teaching Hospital for dogs with extremity bone tumors. Optimal dose delivery to the tumor during SBRT treatment can be limited by uncertainty in skin dose calculation. The aim of thismore » study was to characterize the difference between measured and calculated radiation dose by the Varian Eclipse (Varian Medical Systems, Palo Alto, CA) AAA treatment planning algorithm (for 1-mm, 2-mm, and 5-mm calculation voxel dimensions) as a function of distance from the skin surface. The study used Gafchromic EBT film (International Specialty Products, Wayne, NJ), FilmQA analysis software, a limb phantom constructed from plastic water Trade-Mark-Sign (fluke Biomedical, Everett, WA) and a canine cadaver forelimb. The limb phantom was exposed to 6-MV treatments consisting of a single-beam, a pair of parallel opposed beams, and a 7-beam coplanar treatment plan. The canine forelimb was exposed to the 7-beam coplanar plan. Radiation dose to the forelimb skin at the surface and at depths of 1.65 mm and 1.35 mm below the skin surface were also measured with the Gafchromic film. The calculation algorithm estimated the dose well at depths beyond buildup for all calculation voxel sizes. The calculation algorithm underestimated the dose in portions of the buildup region of tissue for all comparisons, with the most significant differences observed in the 5-mm calculation voxel and the least difference in the 1-mm voxel. Results indicate a significant difference between measured and

  13. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lewis, David F., E-mail: rcfilmconsulting@gmail.com; Chan, Maria F.

    Purpose: The new radiochromic film, GAFChromic EBT-XD, contains the same active material, lithium-10,12-pentacosadiynoate, as GAFChromic EBT3, but the crystalline form is different. This work investigates the effect of this change on the well-known lateral response artifact when EBT-XD film is digitized on a flatbed scanner. Methods: The dose response of a single production lot of EBT-XD was characterized by scanning an unexposed film plus a set of films exposed to doses between 2.5 and 50 Gy using 6 MV photons. To characterize the lateral response artifact, the authors used the unexposed film plus a subset of samples exposed to dosesmore » between 20 and 50 Gy. Digital images of these films were acquired at seven discrete lateral locations perpendicular to the scan direction on three Epson 10000XL scanners. Using measurements at the discrete lateral positions, the scanner responses were determined as a function of the lateral position of the film. From the data for each scanner, a set of coefficients were derived whereby measured response values could be corrected to remove the effects of the lateral response artifact. The EBT-XD data were analyzed as in their previous work and compared to results reported for EBT3 in that paper. Results: For films scanned in the same orientation and having equal responses, the authors found that the lateral response artifact for EBT-XD and EBT3 films was remarkably similar. For both films, the artifact increases with increased net response. However, as EBT-XD is less sensitive than EBT3, a greater exposure dose is required to reach the same net response. On this basis, the lower sensitivity of EBT-XD relative to EBT3 results in less net response change for equal exposure and a reduction in the impact of the lateral response artifact. Conclusions: The shape of the crystalline active component in EBT-XD and EBT3 does not affect the fundamental existence of the lateral response artifact when the films are digitized on flatbed

  14. High throughput film dosimetry in homogeneous and heterogeneous media for a small animal irradiator

    PubMed Central

    Wack, L.; Ngwa, W.; Tryggestad, E.; Tsiamas, P.; Berbeco, R.; Ng, S.K.; Hesser, J.

    2013-01-01

    Purpose We have established a high-throughput Gafchromic film dosimetry protocol for narrow kilo-voltage beams in homogeneous and heterogeneous media for small-animal radiotherapy applications. The kV beam characterization is based on extensive Gafchromic film dosimetry data acquired in homogeneous and heterogeneous media. An empirical model is used for parameterization of depth and off-axis dependence of measured data. Methods We have modified previously published methods of film dosimetry to suit the specific tasks of the study. Unlike film protocols used in previous studies, our protocol employs simultaneous multichannel scanning and analysis of up to nine Gafchromic films per scan. A scanner and background correction were implemented to improve accuracy of the measurements. Measurements were taken in homogeneous and inhomogeneous phantoms at 220 kVp and a field size of 5 × 5 mm2. The results were compared against Monte Carlo simulations. Results Dose differences caused by variations in background signal were effectively removed by the corrections applied. Measurements in homogeneous phantoms were used to empirically characterize beam data in homogeneous and heterogeneous media. Film measurements in inhomogeneous phantoms and their empirical parameterization differed by about 2%–3%. The model differed from MC by about 1% (water, lung) to 7% (bone). Good agreement was found for measured and modelled off-axis ratios. Conclusions EBT2 films are a valuable tool for characterization of narrow kV beams, though care must be taken to eliminate disturbances caused by varying background signals. The usefulness of the empirical beam model in interpretation and parameterization of film data was demonstrated. PMID:23510532

  15. Dosimetry for 131Cs and 125I seeds in solid water phantom using radiochromic EBT film.

    PubMed

    Chiu-Tsao, Sou-Tung; Napoli, John J; Davis, Stephen D; Hanley, Joseph; Rivard, Mark J

    2014-09-01

    To measure the 2D dose distributions with submillimeter resolution for (131)Cs (model CS-1 Rev2) and (125)I (model 6711) seeds in a Solid Water phantom using radiochromic EBT film for radial distances from 0.06cm to 5cm. To determine the TG-43 dosimetry parameters in water by applying Solid Water to liquid water correction factors generated from Monte Carlo simulations. Each film piece was positioned horizontally above and in close contact with a (131)Cs or (125)I seed oriented horizontally in a machined groove at the center of a Solid Water phantom, one film at a time. A total of 74 and 50 films were exposed to the (131)Cs and (125)I seeds, respectively. Different film sizes were utilized to gather data in different distance ranges. The exposure time varied according to the seed air-kerma strength and film size in order to deliver doses in the range covered by the film calibration curve. Small films were exposed for shorter times to assess the near field, while larger films were exposed for longer times in order to assess the far field. For calibration, films were exposed to either 40kV (M40) or 50kV (M50) x-rays in air at 100.0cm SSD with doses ranging from 0.2Gy to 40Gy. All experimental, calibration and background films were scanned at a 0.02cmpixel resolution using a CCD camera-based microdensitometer with a green light source. Data acquisition and scanner uniformity correction were achieved with Microd3 software. Data analysis was performed using ImageJ, FV, IDL and Excel software packages. 2D dose distributions were based on the calibration curve established for 50kV x-rays. The Solid Water to liquid water medium correction was calculated using the MCNP5 Monte Carlo code. Subsequently, the TG-43 dosimetry parameters in liquid water medium were determined. Values for the dose-rate constants using EBT film were 1.069±0.036 and 0.923±0.031cGyU(-1)h(-1) for (131)Cs and (125)I seed, respectively. The corresponding values determined using the Monte Carlo method

  16. Calibration of EBT2 film by the PDD method with scanner non-uniformity correction.

    PubMed

    Chang, Liyun; Chui, Chen-Shou; Ding, Hueisch-Jy; Hwang, Ing-Ming; Ho, Sheng-Yow

    2012-09-21

    The EBT2 film together with a flatbed scanner is a convenient dosimetry QA tool for verification of clinical radiotherapy treatments. However, it suffers from a relatively high degree of uncertainty and a tedious film calibration process for every new lot of films, including cutting the films into several small pieces, exposing with different doses, restoring them back and selecting the proper region of interest (ROI) for each piece for curve fitting. In this work, we present a percentage depth dose (PDD) method that can accurately calibrate the EBT2 film together with the scanner non-uniformity correction and provide an easy way to perform film dosimetry. All films were scanned before and after the irradiation in one of the two homemade 2 mm thick acrylic frames (one portrait and the other landscape), which was located at a fixed position on the scan bed of an Epson 10 000XL scanner. After the pre-irradiated scan, the film was placed parallel to the beam central axis and sandwiched between six polystyrene plates (5 cm thick each), followed by irradiation of a 20 × 20 cm² 6 MV photon beam. Two different beams on times were used on two different films to deliver a dose to the film ranging from 32 to 320 cGy. After the post-irradiated scan, the net optical densities for a total of 235 points on the beam central axis on the films were auto-extracted and compared with the corresponding depth doses that were calculated through the measurement of a 0.6 cc farmer chamber and the related PDD table to perform the curve fitting. The portrait film location was selected for routine calibration, since the central beam axis on the film is parallel to the scanning direction, where non-uniformity correction is not needed (Ferreira et al 2009 Phys. Med. Biol. 54 1073-85). To perform the scanner non-uniformity calibration, the cross-beam profiles of the film were analysed by referencing the measured profiles from a Profiler™. Finally, to verify our method, the films were

  17. Evaluation of a single-scan protocol for radiochromic film dosimetry.

    PubMed

    Shimohigashi, Yoshinobu; Araki, Fujio; Maruyama, Masato; Nakaguchi, Yuji; Kuwahara, Satoshi; Nagasue, Nozomu; Kai, Yudai

    2015-03-08

    The purpose of this study was to evaluate a single-scan protocol using Gafchromic EBT3 film (EBT3) by comparing it with the commonly used 24-hr measurement protocol for radiochromic film dosimetry. Radiochromic film is generally scanned 24 hr after film exposure (24-hr protocol). The single-scan protocol enables measurement results within a short time using only the verification film, one calibration film, and unirradiated film. The single-scan protocol was scanned 30 min after film irradiation. The EBT3 calibration curves were obtained with the multichannel film dosimetry method. The dose verifications for each protocol were performed with the step pattern, pyramid pattern, and clinical treatment plans for intensity-modulated radiation therapy (IMRT). The absolute dose distributions for each protocol were compared with those calculated by the treatment planning system (TPS) using gamma evaluation at 3% and 3 mm. The dose distribution for the single-scan protocol was within 2% of the 24-hr protocol dose distribution. For the step pattern, the absolute dose discrepancies between the TPS for the single-scan and 24-hr protocols were 2.0 ± 1.8 cGy and 1.4 ± 1.2 cGy at the dose plateau, respectively. The pass rates were 96.0% for the single-scan protocol and 95.9% for the 24-hr protocol. Similarly, the dose discrepancies for the pyramid pattern were 3.6 ± 3.5cGy and 2.9 ± 3.3 cGy, respectively, while the pass rates for the pyramid pattern were 95.3% and 96.4%, respectively. The average pass rates for the four IMRT plans were 96.7% ± 1.8% for the single-scan protocol and 97.3% ± 1.4% for the 24-hr protocol. Thus, the single-scan protocol measurement is useful for dose verification of IMRT, based on its accuracy and efficiency.

  18. Skin dose for head and neck cancer patients treated with intensity-modulated radiation therapy(IMRT)

    NASA Astrophysics Data System (ADS)

    Fu, Hsiao-Ju; Li, Chi-Wei; Tsai, Wei-Ta; Chang, Chih-Chia; Tsang, Yuk-Wah

    2017-11-01

    The reliability of thermoluminescent dosimeters (ultrathin TLD) and ISP Gafchromic EBT2 film to measure the surface dose in phantom and the skin dose in head-and-neck patients treated with intensity-modulated radiation therapy technique(IMRT) is the research focus. Seven-field treatment plans with prescribed dose of 180 cGy were performed on Eclipse treatment planning system which utilized pencil beam calculation algorithm(PBC). In calibration tests, the variance coefficient of the ultrathin TLDs were within 3%. The points on the calibration curve of the Gafchromic film was within 1% variation. Five measurements were taken on phantom using ultrathin TLD and EBT2 film respectively. The measured mean surface doses between ultrathin TLD or EBT2 film were within 5% deviation. Skin doses of 6 patients were measured for initial 5 fractions and the mean dose per-fraction was calculated. If the extrapolated doses for 30 fractions were below 4000 cGy, the skin reaction grading observed according to Radiation Therapy Oncology Group (RTOG) was either grade 1 or grade 2. If surface dose exceeded 5000 cGy in 32 fractions, then grade 3 skin reactions were observed.

  19. The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner

    PubMed Central

    Sim, GS; Ng, KH

    2013-01-01

    Radiochromic and radiographic films are widely used for radiation dosimetry due to the advantage of high spatial resolution and two‐dimensional dose measurement. Different types of scanners, including various models of flatbed scanners, have been used as part of the dosimetry readout procedure. This paper focuses on the characterization of the EBT2 film response in combination with a Microtek ScanMaker 9800XL scanner and the subsequent use in the dosimetric verification of a 3D conformal radiotherapy treatment. The film reproducibility and scanner uniformity of the Microtek ScanMaker 9800XL was studied. A three‐field 3D conformal radiotherapy treatment was planned on an anthropomorphic phantom and EBT2 film measurements were carried out to verify the treatment. The interfilm reproducibility was found to be 0.25%. Over a period of three months, the films darkened by 1%. The scanner reproducibility was ± 2% and a nonuniformity was ±1.9% along the direction perpendicular to the scan direction. EBT2 measurements showed an underdose of 6.2% at high‐dose region compared to TPS predicted dose. This may be due to the inability of the treatment planning system to predict the correct dose distribution in the presence of tissue inhomogeneities and the uncertainty of the scanner reproducibility and uniformity. The use of EBT2 film in conjunction with the axial CT image of the anthropomorphic phantom allows the evaluation of the anatomical location of dose discrepancies between the EBT2 measured dose distribution and TPS predicted dose distribution. PACS number: 87.55.Qr PMID:23835383

  20. SU-G-BRB-08: Investigation On the Magnetic Field Effect On TLDs, OSLDs, and Gafchromic Films Using An MR-Linac

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wen, Z; Wang, J; O’Brien, D

    Purpose: To investigate whether a strong magnetic field (B=1.5 T) can affect dose responses of thermoluminescent dosimeters (TLDs), optically stimulated luminescence dosimeters (OSLDs) and Gafchromic films using an MR-Linac (Elekta) before and after the magnet was ramped down from 1.5 T to 0 T. Methods: Three types of dosimeters (TLDs, OSLDs, EBT3 films) were divided into two groups. Group 1 was first irradiated in a phantom of Solid Water slabs (Standard Imaging) inside a B=1.5 T field with a 7 MV beam from an MR-Linac system. The radiation output at the location of the dosimeters (isocenter at 10 cm depth)more » was measured using an ion chamber (NE2571, Phoenix Dosimetry). Three doses (150, 300, 600 MU, corresponding to 1.18, 2.36, and 4.74 Gy) were delivered to the dosimeters. A week later the MR magnet was ramped down to zero field and dosimeters in Group 2 were irradiated with the same MUs. Dosimeters of each type were read out during the same session (about 4 weeks post irradiation in the B field, and 3 weeks with no B field). The ratios of signals between Group 1 and Group 2 were calculated. Results: Radiation output measured with the chamber was within 1% before and after ramping down the MR magnet. For TLDs, the ratio of signals with B field to signals without B field averaged over three dose levels was 1.003±0.016; for OSLDs, the ratio was 0.994±0.022; for films, the ratios of two batches (different manufacturing dates) were 0.997 and 0.985. Conclusion: Dose responses of all three dosimeters seem not affected by the presence of a 1.5 T magnetic field within uncertainty of ∼2%. More measurements will be conducted to test reproducibility. We acknowledge research support from Elekta AB.« less

  1. Characterization of the proton beam from an IBA Cyclone 18/9 with radiochromic film EBT2

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sansaloni, F.; Lagares, J. I.; Arce, P.

    2012-12-19

    The use of radiochromic films is widespread in different areas of medical physics like radiotherapy and hadrontherapy; however, radiochromic films have been scarcely used in the characterization of proton or deuteron beams generated in biomedical cyclotrons. In this paper the radiochromic film EBT2 was used to study the beam size and the proton beam energy of an IBA Cyclone 18/9 cyclotron. The results indicate that the beam size can be easily measured at a very low expense; however, an accurate determination of the beam energy might require the implementation of certain experimental improvements.

  2. Comparison of dose response functions for EBT3 model GafChromic™ film dosimetry system.

    PubMed

    Aldelaijan, Saad; Devic, Slobodan

    2018-05-01

    Different dose response functions of EBT3 model GafChromic™ film dosimetry system have been compared in terms of sensitivity as well as uncertainty vs. error analysis. We also made an assessment of the necessity of scanning film pieces before and after irradiation. Pieces of EBT3 film model were irradiated to different dose values in Solid Water (SW) phantom. Based on images scanned in both reflection and transmission mode before and after irradiation, twelve different response functions were calculated. For every response function, a reference radiochromic film dosimetry system was established by generating calibration curve and by performing the error vs. uncertainty analysis. Response functions using pixel values from the green channel demonstrated the highest sensitivity in both transmission and reflection mode. All functions were successfully fitted with rational functional form, and provided an overall one-sigma uncertainty of better than 2% for doses above 2 Gy. Use of pre-scanned images to calculate response functions resulted in negligible improvement in dose measurement accuracy. Although reflection scanning mode provides higher sensitivity and could lead to a more widespread use of radiochromic film dosimetry, it has fairly limited dose range and slightly increased uncertainty when compared to transmission scan based response functions. Double-scanning technique, either in transmission or reflection mode, shows negligible improvement in dose accuracy as well as a negligible increase in dose uncertainty. Normalized pixel value of the images scanned in transmission mode shows linear response in a dose range of up to 11 Gy. Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  3. Evaluation of a single‐scan protocol for radiochromic film dosimetry

    PubMed Central

    Araki, Fujio; Maruyama, Masato; Nakaguchi, Yuji; Kuwahara, Satoshi; Nagasue, Nozomu; Kai, Yudai

    2015-01-01

    The purpose of this study was to evaluate a single‐scan protocol using Gafchromic EBT3 film (EBT3) by comparing it with the commonly used 24‐hr measurement protocol for radiochromic film dosimetry. Radiochromic film is generally scanned 24 hr after film exposure (24‐hr protocol). The single‐scan protocol enables measurement results within a short time using only the verification film, one calibration film, and unirradiated film. The single‐scan protocol was scanned 30 min after film irradiation. The EBT3 calibration curves were obtained with the multichannel film dosimetry method. The dose verifications for each protocol were performed with the step pattern, pyramid pattern, and clinical treatment plans for intensity‐modulated radiation therapy (IMRT). The absolute dose distributions for each protocol were compared with those calculated by the treatment planning system (TPS) using gamma evaluation at 3% and 3 mm. The dose distribution for the single‐scan protocol was within 2% of the 24‐hr protocol dose distribution. For the step pattern, the absolute dose discrepancies between the TPS for the single‐scan and 24‐hr protocols were 2.0±1.8 cGy and 1.4±1.2 cGy at the dose plateau, respectively. The pass rates were 96.0% for the single‐scan protocol and 95.9% for the 24‐hr protocol. Similarly, the dose discrepancies for the pyramid pattern were 3.6±3.5 cGy and 2.9±3.3 cGy, respectively, while the pass rates for the pyramid pattern were 95.3% and 96.4%, respectively. The average pass rates for the four IMRT plans were 96.7%±1.8% for the single‐scan protocol and 97.3%±1.4% for the 24‐hr protocol. Thus, the single‐scan protocol measurement is useful for dose verification of IMRT, based on its accuracy and efficiency. PACS number: 87.55.Qr PMID:26103194

  4. Assessment of peak skin dose in interventional cardiology: A comparison between Gafchromic film and dosimetric software em.dose.

    PubMed

    Greffier, J; Van Ngoc Ty, C; Bonniaud, G; Moliner, G; Ledermann, B; Schmutz, L; Cornillet, L; Cayla, G; Beregi, J P; Pereira, F

    2017-06-01

    To compare the use of a dose mapping software to Gafchromic film measurement for a simplified peak skin dose (PSD) estimation in interventional cardiology procedure. The study was conducted on a total of 40 cardiac procedures (20 complex coronary angioplasty of chronic total occlusion (CTO) and 20 coronary angiography and coronary angioplasty (CA-PTCA)) conducted between January 2014 to December 2015. PSD measurement (PSD Film ) was obtained by placing XR-RV3 Gafchromic under the patient's back for each procedure. PSD (PSD em.dose ) was computed with the software em.dose©. The calculation was performed on the dose metrics collected from the private dose report of each procedure. Two calculation methods (method A: fluoroscopic kerma equally spread on cine acquisition and B: fluoroscopic kerma is added to one air Kerma cine acquisition that contributes to the PSD) were used to calculate the fluoroscopic dose contribution as fluoroscopic data were not recorded in our interventional room. Statistical analyses were carried out to compare PSD Film and PSD em.dose . The PSD Film median (1st quartile; 3rd quartile) was 0.251(0.190;0.336)Gy for CA-PTCA and 1.453(0.767;2.011)Gy for CTO. For method-A, the PSD em.dose was 0.248(0.182;0.369)Gy for CA-PTCA and 1.601(0.892;2.178)Gy for CTO, and 0.267(0.223;0.446)Gy and 1.75 (0.912;2.584)Gy for method-B, respectively. For the two methods, the correlation between PSD Film and PSD em.dose was strong. For all cardiology procedures investigated, the mean deviation between PSD Film and PSD em.dose was 3.4±21.1% for method-A and 17.3%±23.9% for method-B. The dose mapping software is convenient to calculate peak skin dose in interventional cardiology. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  5. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lynch, Bart D.; Kozelka, Jakub; Ranade, Manisha K.

    In this study, we present three significant artifacts that have the potential to negatively impact the accuracy and precision of film dosimetry measurements made using GAFCHROMIC[reg] EBT radiochromic film when read out with CCD flatbed scanners. Films were scanned using three commonly employed instruments: a Macbeth TD932 spot densitometer, an Epson Expression 1680 CCD array scanner, and a Microtek ScanMaker i900 CCD array scanner. For the two scanners we assessed the variation in optical density (OD) of GAFCHROMIC EBT film with scanning bed position, angular rotation of the film with respect to the scan line direction, and temperature inside themore » scanner due to repeated scanning. Scanning uniform radiochromic films demonstrated a distinct bowing effect in profiles in the direction of the CCD array with a nonuniformity of up to 17%. Profiles along a direction orthogonal to the CCD array demonstrated a 7% variation. A strong angular dependence was found in measurements made with the flatbed scanners; the effect could not be reproduced with the spot densitometer. An IMRT quality assurance film was scanned twice rotating the film 90 deg. between the scans. For films scanned on the Epson scanner, up to 12% variation was observed in unirradiated EBT films rotated between 0 deg. and 90 deg. , which decreased to approximately 8% for EBT films irradiated to 300 cGy. Variations of up to 80% were observed for films scanned with the Microtek scanner. The scanners were found to significantly increase the film temperature with repeated scanning. Film temperature between 18 and 33 deg. C caused OD changes of approximately 7%. Considering these effects, we recommend adherence to a strict scanning protocol that includes: maintaining the orientation of films scanned on flatbed scanners, limiting scanning to the central portion of the scanner bed, and limiting the number of consecutive scans to minimize changes in OD caused by film heating.« less

  6. SU-E-T-519: Investigation of the CyberKnife MultiPlan Monte Carlo Dose Calculation Using EBT3 Film Absolute Dosimetry for Delivery in a Heterogeneous Thorax Phantom

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lamberto, M; Chen, H; Huang, K

    2015-06-15

    Purpose To characterize the Cyberknife (CK) robotic system’s dosimetric accuracy of the delivery of MultiPlan’s Monte Carlo dose calculations using EBT3 radiochromic film inserted in a thorax phantom. Methods The CIRS XSight Lung Tracking (XLT) Phantom (model 10823) was used in this study with custom cut EBT3 film inserted in the horizontal (coronal) plane inside the lung tissue equivalent phantom. CK MultiPlan v3.5.3 with Monte Carlo dose calculation algorithm (1.5 mm grid size, 2% statistical uncertainty) was used to calculate a clinical plan for a 25-mm lung tumor lesion, as contoured by the physician, and then imported onto the XLTmore » phantom CT. Using the same film batch, the net OD to dose calibration curve was obtained using CK with the 60 mm fixed cone by delivering 0– 800 cGy. The test films (n=3) were irradiated using 325 cGy to the prescription point. Films were scanned 48 hours after irradiation using an Epson v700 scanner (48 bits color scan, extracted red channel only, 96 dpi). Percent absolute dose and relative isodose distribution difference relative to the planned dose were quantified using an in-house QA software program. Multiplan Monte Carlo dose calculation was validated using RCF dosimetry (EBT3) and gamma index criteria of 3%/3mm and 2%/2mm for absolute dose and relative isodose distribution measurement comparisons. Results EBT3 film measurements of the patient plans calculated with Monte Carlo in MultiPlan resulted in an absolute dose passing rate of 99.6±0.4% for the Gamma Index of 3%/3mm, 10% dose threshold, and 95.6±4.4% for 2%/2mm, 10% threshold criteria. The measured central axis absolute dose was within 1.2% (329.0±2.5 cGy) of the Monte Carlo planned dose (325.0±6.5 cGy) for that same point. Conclusion MultiPlan’s Monte Carlo dose calculation was validated using the EBT3 film absolute dosimetry for delivery in a heterogeneous thorax phantom.« less

  7. In vivo dosimetry using Gafchromic films during pelvic intraoperative electron radiation therapy (IOERT)

    PubMed Central

    Costa, Filipa; Gomes, Dora; Magalhães, Helena; Arrais, Rosário; Moreira, Graciete; Cruz, Maria Fátima; Silva, José Pedro; Santos, Lúcio; Sousa, Olga

    2016-01-01

    Objective: To characterize in vivo dose distributions during pelvic intraoperative electron radiation therapy (IOERT) for rectal cancer and to assess the alterations introduced by irregular irradiation surfaces in the presence of bevelled applicators. Methods: In vivo measurements were performed with Gafchromic films during 32 IOERT procedures. 1 film per procedure was used for the first 20 procedures. The methodology was then optimized for the remaining 12 procedures by using a set of 3 films. Both the average dose and two-dimensional dose distributions for each film were determined. Phantom measurements were performed for comparison. Results: For flat and concave surfaces, the doses measured in vivo agree with expected values. For concave surfaces with step-like irregularities, measured doses tend to be higher than expected doses. Results obtained with three films per procedure show a large variability along the irradiated surface, with important differences from expected profiles. These results are consistent with the presence of surface hotspots, such as those observed in phantoms in the presence of step-like irregularities, as well as fluid build-up. Conclusion: Clinical dose distributions in the IOERT of rectal cancer are often different from the references used for prescription. Further studies are necessary to assess the impact of these differences on treatment outcomes. In vivo measurements are important, but need to be accompanied by accurate imaging of positioning and irradiated surfaces. Advances in knowledge: These results confirm that surface irregularities occur frequently in rectal cancer IOERT and have a measurable effect on the dose distribution. PMID:27188847

  8. On the calibration process of film dosimetry: OLS inverse regression versus WLS inverse prediction.

    PubMed

    Crop, F; Van Rompaye, B; Paelinck, L; Vakaet, L; Thierens, H; De Wagter, C

    2008-07-21

    The purpose of this study was both putting forward a statistically correct model for film calibration and the optimization of this process. A reliable calibration is needed in order to perform accurate reference dosimetry with radiographic (Gafchromic) film. Sometimes, an ordinary least squares simple linear (in the parameters) regression is applied to the dose-optical-density (OD) curve with the dose as a function of OD (inverse regression) or sometimes OD as a function of dose (inverse prediction). The application of a simple linear regression fit is an invalid method because heteroscedasticity of the data is not taken into account. This could lead to erroneous results originating from the calibration process itself and thus to a lower accuracy. In this work, we compare the ordinary least squares (OLS) inverse regression method with the correct weighted least squares (WLS) inverse prediction method to create calibration curves. We found that the OLS inverse regression method could lead to a prediction bias of up to 7.3 cGy at 300 cGy and total prediction errors of 3% or more for Gafchromic EBT film. Application of the WLS inverse prediction method resulted in a maximum prediction bias of 1.4 cGy and total prediction errors below 2% in a 0-400 cGy range. We developed a Monte-Carlo-based process to optimize calibrations, depending on the needs of the experiment. This type of thorough analysis can lead to a higher accuracy for film dosimetry.

  9. Characterization of Al2O3 optically stimulated luminescence films for 2D dosimetry using a 6 MV photon beam

    NASA Astrophysics Data System (ADS)

    Ahmed, M. F.; Shrestha, N.; Schnell, E.; Ahmad, S.; Akselrod, M. S.; Yukihara, E. G.

    2016-11-01

    This work evaluates the dosimetric properties of newly developed optically stimulated luminescence (OSL) films, fabricated with either Al2O3:C or Al2O3:C,Mg, using a prototype laser scanning reader, a developed image reconstruction algorithm, and a 6 MV therapeutic photon beam. Packages containing OSL films (Al2O3:C and Al2O3:C,Mg) and a radiochromic film (Gafchromic EBT3) were irradiated using a 6 MV photon beam using different doses, field sizes, with and without wedge filter. Dependence on film orientation of the OSL system was also tested. Diode-array (MapCHECK) and ionization chamber measurements were performed for comparison. The OSLD film doses agreed with the MapCHECK and ionization chamber data within the experimental uncertainties (<2% at 1.5 Gy). The system background and minimum detectable dose (MDD) were  <0.5 mGy, and the dose response was approximately linear from the MDD up to a few grays (the linearity correction was  <10% up to ~2-4 Gy), with no saturation up to 30 Gy. The dose profiles agreed with those obtained using EBT3 films (analyzed using the triple channel method) in the high dose regions of the images. In the low dose regions, the dose profiles from the OSLD films were more reproducible than those from the EBT3 films. We also demonstrated that the OSL film data are independent on scan orientation and field size over the investigated range. The results demonstrate the potential of OSLD films for 2D dosimetry, particularly for the characterization of small fields, due to their wide dynamic range, linear response, resolution and dosimetric properties. The negligible background and potential simple calibration make these OSLD films suitable for remote audits. The characterization presented here may motivate further commercial development of a 2D dosimetry system based on the OSL from Al2O3:C or Al2O3:C,Mg.

  10. I-125 ROPES eye plaque dosimetry: validation of a commercial 3D ophthalmic brachytherapy treatment planning system and independent dose calculation software with GafChromic® EBT3 films.

    PubMed

    Poder, Joel; Corde, Stéphanie

    2013-12-01

    The purpose of this study was to measure the dose distributions for different Radiation Oncology Physics and Engineering Services, Australia (ROPES) type eye plaques loaded with I-125 (model 6711) seeds using GafChromic(®) EBT3 films, in order to verify the dose distributions in the Plaque Simulator™ (PS) ophthalmic 3D treatment planning system. The brachytherapy module of RADCALC(®) was used to independently check the dose distributions calculated by PS. Correction factors were derived from the measured data to be used in PS to account for the effect of the stainless steel ROPES plaque backing on the 3D dose distribution. Using GafChromic(®) EBT3 films inserted in a specially designed Solid Water™ eye ball phantom, dose distributions were measured three-dimensionally both along and perpendicular to I-125 (model 6711) loaded ROPES eye plaque's central axis (CAX) with 2 mm depth increments. Each measurement was performed in full scatter conditions both with and without the stainless steel plaque backing attached to the eye plaque, to assess its effect on the dose distributions. Results were compared to the dose distributions calculated by Plaque Simulator™ and checked independently with RADCALC(®). The EBT3 film measurements without the stainless steel backing were found to agree with PS and RADCALC(®) to within 2% and 4%, respectively, on the plaque CAX. Also, RADCALC(®) was found to agree with PS to within 2%. The CAX depth doses measured using EBT3 film with the stainless steel backing were observed to result in a 4% decrease relative to when the backing was not present. Within experimental uncertainty, the 4% decrease was found to be constant with depth and independent of plaque size. Using a constant dose correction factor of T = 0.96 in PS, where the calculated dose for the full water scattering medium is reduced by 4% in every voxel in the dose grid, the effect of the plaque backing was accurately modeled in the planning system. Off-axis profiles

  11. Gafchromic EBT‐XD film: Dosimetry characterization in high‐dose, volumetric‐modulated arc therapy

    PubMed Central

    Ozawa, Shuichi; Hosono, Fumika; Sumida, Naoki; Okazue, Toshiya; Yamada, Kiyoshi; Nagata, Yasushi

    2016-01-01

    Radiochromic films are important tools for assessing complex dose distributions. Gafchromic EBT‐XD films have been designed for optimal performance in the 40–4,000 cGy dose range. We investigated the dosimetric characteristics of these films, including their dose‐response, postexposure density growth, and dependence on scanner orientation, beam energy, and dose rate with applications to high‐dose volumetric‐modulated arc therapy (VMAT) verification. A 10 MV beam from a TrueBeam STx linear accelerator was used to irradiate the films with doses in the 0–4,000 cGy range. Postexposure coloration was analyzed at postirradiation times ranging from several minutes to 48 h. The films were also irradiated with 6 MV (dose rate (DR): 600 MU/min), 6 MV flattening filter‐free (FFF) (DR: 1,400 MU/ min), and 10 MV FFF (DR: 2,400 MU/min) beams to determine the energy and dose‐rate dependence. For clinical examinations, we compared the dose distribution measured with EBT‐XD films and calculated by the planning system for four VMAT cases. The red channel of the EBT‐XD film exhibited a wider dynamic range than the green and blue channels. Scanner orientation yielded a variation of ∼3% in the net optical density (OD). The difference between the film front and back scan orientations was negligible, with variation of ∼1.3% in the net OD. The net OD increased sharply within the first 6 hrs after irradiation and gradually afterwards. No significant difference was observed for the beam energy and dose rate, with a variation of ∼1.5% in the net OD. The gamma passing rates (at 3%, 3 mm) between the film‐ measured and treatment planning system (TPS)‐calculated dose distributions under a high dose VMAT plan in the absolute dose mode were more than 98.9%. PACS number(s): 87.56 Fc PMID:27929504

  12. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Devic, Slobodan; Tomic, Nada; Aldelaijan, Saad

    Purpose: Despite numerous advantages of radiochromic film dosimeter (high spatial resolution, near tissue equivalence, low energy dependence) to measure a relative dose distribution with film, one needs to first measure an absolute dose (following previously established reference dosimetry protocol) and then convert measured absolute dose values into relative doses. In this work, we present result of our efforts to obtain a functional form that would linearize the inherently nonlinear dose-response curve of the radiochromic film dosimetry system. Methods: Functional form [{zeta}= (-1){center_dot}netOD{sup (2/3)}/ln(netOD)] was derived from calibration curves of various previously established radiochromic film dosimetry systems. In order to testmore » the invariance of the proposed functional form with respect to the film model used we tested it with three different GAFCHROMIC Trade-Mark-Sign film models (EBT, EBT2, and EBT3) irradiated to various doses and scanned on a same scanner. For one of the film models (EBT2), we tested the invariance of the functional form to the scanner model used by scanning irradiated film pieces with three different flatbed scanner models (Epson V700, 1680, and 10000XL). To test our hypothesis that the proposed functional argument linearizes the response of the radiochromic film dosimetry system, verification tests have been performed in clinical applications: percent depth dose measurements, IMRT quality assurance (QA), and brachytherapy QA. Results: Obtained R{sup 2} values indicate that the choice of the functional form of the new argument appropriately linearizes the dose response of the radiochromic film dosimetry system we used. The linear behavior was insensitive to both film model and flatbed scanner model used. Measured PDD values using the green channel response of the GAFCHROMIC Trade-Mark-Sign EBT3 film model are well within {+-}2% window of the local relative dose value when compared to the tabulated Cobalt-60 data. It

  13. Estimation of Eye Lens Dose During Brain Scans Using Gafchromic Xr-QA2 Film in Various Multidetector CT Scanners.

    PubMed

    Akhilesh, Philomina; Kulkarni, Arti R; Jamhale, Shramika H; Sharma, S D; Kumar, Rajesh; Datta, D

    2017-04-25

    The purpose of this study was to estimate eye lens dose during brain scans in 16-, 64-, 128- and 256-slice multidetector computed tomography (CT) scanners in helical acquisition mode and to test the feasibility of using radiochromic film as eye lens dosemeter during CT scanning. Eye lens dose measurements were performed using Gafchromic XR-QA2 film on a polystyrene head phantom designed with outer dimensions equivalent to the head size of a reference Indian man. The response accuracy of XR-QA2 film was validated by using thermoluminescence dosemeters. The eye lens dose measured using XR-QA2 film on head phantom for plain brain scanning in helical mode ranged from 43.8 to 45.8 mGy. The XR-QA2 film measured dose values were in agreement with TLD measured dose values within a maximum variation of 8.9%. The good correlation between the two data sets confirms the viability of using XR-QA2 film for eye lens dosimetry. © The Author 2016. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  14. TH-CD-BRA-04: Effect of a Strong Magnetic Field On TLDs, OSLDs, and Gafchromic Films Using An Electromagnet

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wang, J; Rubinstein, A; Ohrt, J

    Purpose: To study the effect of strong magnetic field on three types of dosimeters using an electromagnet inside a Linac vault. Materials and Methods: Three types of dosimeters, thermoluminescent Dosimeters (TLDs), optically stimulated luminescent Dosimeters (OSLDs), and EBT3 Film were used to measure radiation dose response inside an electromagnet that could produce a strong magnetic field (B>1.5 T). The dosimeters were placed inside a plastic phantom between the two poles of the magnet, at approximately 3 meters from the iso-center of an Elekta Versa HD Linac. The B field was calibrated with a Gauss meter (Model: GM-2, AlphaLab Inc). Themore » dosimeters received ∼2 Gy with and without the presence of the 1.5 T magnetic field. The EBT3 films were scanned 24 hours before and 24 hours after irradiation. The TLD dosimeters were read 1 week after irradiation. The OSLDs were read two weeks after irradiation. The ratios of signals of dosimeters irradiated with the B field to the signals without the B field were calculated. Two experiments have been conducted so far. Results: The ratios (averaged over two experiments) of dosimeter signals with vs without B field were 0.994 for films, 0.994 for OSLDs, and 1.002 for TLDs. The statistical uncertainty was ∼3%. Conclusions: The three types of dosimeters (film, TLD, OSLD) seem not affected by the presence of a magnetic field (B=1.5 T) with the uncertainty of ∼3%. They may be suitable for QA purposes in a strong B field up to 1.5 T. More measurements will be conducted for reproducibility testing. We acknowledge research support from Elekta AB.« less

  15. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Battum, LJ van; Heukelom, S

    Purpose This study investigates the origin of lateral optical density (OD) variation for Gafchromic film (EBT and EBT2) scanned in transmission mode with Epson flatbed scanners (1680 Expression Pro and 10000XL). Effects investigated are: cross talk, optical path length and polarization. Methods Cross talk has been examined with triangular shaped light-transmission sheets with OD ranging from 0 to opaque. Optical path length has been studied with absorptive and reflective OD-filters (OD range 0.2 to 2.0). Dependency on light-polarization on the scanner read out has been investigated using linear polarizer sheets. All experiments have been performed at centre scanner position (normmore » point) and at several lateral scan positions, without and with (un)irradiated EBT-film. Dose values used ranged between 0.2 to 9 Gy, yielding an OD-range between 0.25 to 1.1. Results The lateral OD variation is dose dependent and increases up to 14% at most lateral position for dose up to 9 Gy. Cross talk effect contributes to 0.5% in clinical used OD ranges but equals 2% for extreme high dose gradients. Film induced optical path length will effect the lateral OD variation up to 3% at most lateral points. Light polarization is inherent present in these scanners due to multiple reflection on mirrors. In addition film induced polarization is the most important effect generating the observed lateral OD variation. Both Gafchromic film base and sensitive layer have polarizing capabilities; for the sensitive layer its influence is dose dependent. Conclusions Lateral OD variation origins from optical physics (i.e. polarization and reflection) related to scanner and film construction. Cross talk can be ignored in film dosimetry for clinical used dose values and gradients. Therefore it is recommended to determine the lateral OD variation per film type and scanner.« less

  16. I-125 ROPES eye plaque dosimetry: Validation of a commercial 3D ophthalmic brachytherapy treatment planning system and independent dose calculation software with GafChromic{sup ®} EBT3 films

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Poder, Joel; Corde, Stéphanie

    Purpose: The purpose of this study was to measure the dose distributions for different Radiation Oncology Physics and Engineering Services, Australia (ROPES) type eye plaques loaded with I-125 (model 6711) seeds using GafChromic{sup ®} EBT3 films, in order to verify the dose distributions in the Plaque Simulator™ (PS) ophthalmic 3D treatment planning system. The brachytherapy module of RADCALC{sup ®} was used to independently check the dose distributions calculated by PS. Correction factors were derived from the measured data to be used in PS to account for the effect of the stainless steel ROPES plaque backing on the 3D dose distribution.Methods:more » Using GafChromic{sup ®} EBT3 films inserted in a specially designed Solid Water™ eye ball phantom, dose distributions were measured three-dimensionally both along and perpendicular to I-125 (model 6711) loaded ROPES eye plaque's central axis (CAX) with 2 mm depth increments. Each measurement was performed in full scatter conditions both with and without the stainless steel plaque backing attached to the eye plaque, to assess its effect on the dose distributions. Results were compared to the dose distributions calculated by Plaque Simulator™ and checked independently with RADCALC{sup ®}.Results: The EBT3 film measurements without the stainless steel backing were found to agree with PS and RADCALC{sup ®} to within 2% and 4%, respectively, on the plaque CAX. Also, RADCALC{sup ®} was found to agree with PS to within 2%. The CAX depth doses measured using EBT3 film with the stainless steel backing were observed to result in a 4% decrease relative to when the backing was not present. Within experimental uncertainty, the 4% decrease was found to be constant with depth and independent of plaque size. Using a constant dose correction factor of T= 0.96 in PS, where the calculated dose for the full water scattering medium is reduced by 4% in every voxel in the dose grid, the effect of the plaque backing was

  17. Multidimensional dosimetry of {sup 106}Ru eye plaques using EBT3 films and its impact on treatment planning

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Heilemann, G., E-mail: gerd.heilemann@meduniwien.ac.at; Kostiukhina, N.; Nesvacil, N.

    2015-10-15

    Purpose: The purpose of this study was to establish a method to perform multidimensional radiochromic film measurements of {sup 106}Ru plaques and to benchmark the resulting dose distributions against Monte Carlo simulations (MC), microdiamond, and diode measurements. Methods: Absolute dose rates and relative dose distributions in multiple planes were determined for three different plaque models (CCB, CCA, and COB), and three different plaques per model, using EBT3 films in an in-house developed polystyrene phantom and the MCNP6 MC code. Dose difference maps were generated to analyze interplaque variations for a specific type, and for comparing measurements against MC simulations. Furthermore,more » dose distributions were validated against values specified by the manufacturer (BEBIG) and microdiamond and diode measurements in a water scanning phantom. Radial profiles were assessed and used to estimate dosimetric margins for a given combination of representative tumor geometry and plaque size. Results: Absolute dose rates at a reference depth of 2 mm on the central axis of the plaque show an agreement better than 5% (10%) when comparing film measurements (MCNP6) to the manufacturer’s data. The reproducibility of depth-dose profile measurements was <7% (2 SD) for all investigated detectors and plaque types. Dose difference maps revealed minor interplaque deviations for a specific plaque type due to inhomogeneities of the active layer. The evaluation of dosimetric margins showed that for a majority of the investigated cases, the tumor was not completely covered by the 100% isodose prescribed to the tumor apex if the difference between geometrical plaque size and tumor base ≤4 mm. Conclusions: EBT3 film dosimetry in an in-house developed phantom was successfully used to characterize the dosimetric properties of different {sup 106}Ru plaque models. The film measurements were validated against MC calculations and other experimental methods and showed a good

  18. WE-AB-BRB-08: Progress Towards a 2D OSL Dosimetry System Using Al2O3:C Films

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ahmed, M F; Yukihara, E; Schnell, E

    Purpose: To develop a 2D dosimetry system based on the optically stimulated luminescence (OSL) of Al{sub 2}O{sub 3}:C films for medical applications. Methods: A 2D laser scanning OSL reader was built for readout of newly developed Al2O3:C films (Landauer Inc.). An image reconstruction algorithm was developed to correct for inherent effects introduced by reader design and detector properties. The system was tested using irradiations with photon and carbon ion beams. A calibration was obtained using a 6 MV photon beam from clinical accelerator and the dose measurement precision was tested using a range of doses and different dose distributions (flatmore » field and wedge field). The dynamic range and performance of the system in the presence of large dose gradients was also tested using 430 MeV/u {sup 12}C single and multiple pencil beams. All irradiations were performed with Gafchromic EBT3 film for comparison. Results: Preliminary results demonstrate a near-linear OSL dose response to photon fields and the ability to measure dose in dose distributions such as flat field and wedge field. Tests using {sup 12}C pencil beam demonstrate ability to measure doses over four orders of magnitude. The dose profiles measured by the OSL film generally agreed well with that measured by the EBT3 film. The OSL image signal-to-noise ratio obtained in the current conditions require further improvement. On the other hand, EBT3 films had large uncertainties in the low dose region due to film-to-film or intra-film variation in the background. Conclusion: A 2D OSL dosimetry system was developed and initial tests have demonstrated a wide dynamic range as well as good agreement between the delivered and measured doses. The low background, wide dynamic range and wide range of linearity in dose response observed for the Al{sub 2}O{sub 3}:C OSL film can be beneficial for dosimetry in radiation therapy applications, especially for small field dosimetry. This work has been funded by Landauer

  19. Film dosimetry using a smart device camera: a feasibility study for point dose measurements

    NASA Astrophysics Data System (ADS)

    Aland, Trent; Jhala, Ekta; Kairn, Tanya; Trapp, Jamie

    2017-10-01

    In this work, a methodology for using a smartphone camera, in conjunction with a light-tight box operating in reflective transmission mode, is investigated as a proof of concept for use as a film dosimetry system. An imaging system was designed to allow the camera of a smartphone to be used as a pseudo densitometer. Ten pieces of Gafchromic EBT3 film were irradiated to doses up to 16.89 Gy and used to evaluate the effects of reproducibility and orientation, as well as the ability to create an accurate dose response curve for the smartphone based dosimetry system, using all three colour channels. Results were compared to a flatbed scanner system. Overall uncertainty was found to be best for the red channel with an uncertainty of 2.4% identified for film irradiated to 2.5 Gy and digitised using the smartphone system. This proof of concept exercise showed that although uncertainties still exceed a flatbed scanner system, the smartphone system may be useful for providing point dose measurements in situations where conventional flatbed scanners (or other dosimetry systems) are unavailable or unaffordable.

  20. Film dosimetry using a smart device camera: a feasibility study for point dose measurements.

    PubMed

    Aland, Trent; Jhala, Ekta; Kairn, Tanya; Trapp, Jamie

    2017-10-03

    In this work, a methodology for using a smartphone camera, in conjunction with a light-tight box operating in reflective transmission mode, is investigated as a proof of concept for use as a film dosimetry system. An imaging system was designed to allow the camera of a smartphone to be used as a pseudo densitometer. Ten pieces of Gafchromic EBT3 film were irradiated to doses up to 16.89 Gy and used to evaluate the effects of reproducibility and orientation, as well as the ability to create an accurate dose response curve for the smartphone based dosimetry system, using all three colour channels. Results were compared to a flatbed scanner system. Overall uncertainty was found to be best for the red channel with an uncertainty of 2.4% identified for film irradiated to 2.5 Gy and digitised using the smartphone system. This proof of concept exercise showed that although uncertainties still exceed a flatbed scanner system, the smartphone system may be useful for providing point dose measurements in situations where conventional flatbed scanners (or other dosimetry systems) are unavailable or unaffordable.

  1. SU-E-T-275: Dose Build Up and Bolusing Characteristics for Total Body Irradiation Dosimetry

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Butson, M; Pope, D; Whitaker, M

    2015-06-15

    Purpose: Total Body Irradiation (TBI) treatments are mainly used in a preparative regimen for haematopoietic stem cell (or bone marrow) transplantation. Our standard regimen is a 12 Gy / 6 fraction bi-daily technique. To evaluate the delivered dose homogeneity to the patient, EBT3 Gafchromic film is positioned at the head, neck, chest, pelvis and groin for all fractions. This work investigates and quantifies the build-up dose characteristics at TBI distances and requirements for in-vivo dosimetry bolusing. Methods: Percentage dose build up characteristics of photon beams have been investigated at large extended SSD’s using parallel plate ionisations chambers (Attix) and EBT3more » Gafchromic film. Measurements were made to open fields at different field sizes as well as large 40cm × 40cm fields with differing scatter conditions such as the introduction of standard Perspex scattering plates at different distances to the measurement point. Results: Percentage surface dose measured values for open fields at 300 cm SSD were found to range from 20 % up to 65.5 % for fields of 5 cm × 5 cm to 40 cm × 40 cm. With the introduction of 1cm Perspex scattering plates used in TBI treatments the surface dose values increased up to 83% to 90%, depending on the position of the Perspex scattering plate compared to the measurement point. Our work showed that at least 3mm water equivalent bolus / scatter material should be placed over the EBT3 for accurate dose assessment for TBI treatments. Conclusion: Build up dose characteristics exist at long (300cm) SSD’s including treatments using Perspex scattering plates placed at various distances form the patient during TBI treatment. Top accurately assess the applied dose during treatment, in-vivo dosimeters such as Gafchromic EBT3 should have at least 3mm bolus / scatter material placed over them to measure actual applied doses.« less

  2. Dosimetry for Small Fields in Stereotactic Radiosurgery Using Gafchromic MD-V2-55 Film, TLD-100 and Alanine Dosimeters

    PubMed Central

    Massillon-JL, Guerda; Cueva-Prócel, Diego; Díaz-Aguirre, Porfirio; Rodríguez-Ponce, Miguel; Herrera-Martínez, Flor

    2013-01-01

    This work investigated the suitability of passive dosimeters for reference dosimetry in small fields with acceptable accuracy. Absorbed dose to water rate was determined in nine small radiation fields with diameters between 4 and 35 mm in a Leksell Gamma Knife (LGK) and a modified linear accelerator (linac) for stereotactic radiosurgery treatments. Measurements were made using Gafchromic film (MD-V2-55), alanine and thermoluminescent (TLD-100) dosimeters and compared with conventional dosimetry systems. Detectors were calibrated in terms of absorbed dose to water in 60Co gamma-ray and 6 MV x-ray reference (10×10 cm2) fields using an ionization chamber calibrated at a standards laboratory. Absorbed dose to water rate computed with MD-V2-55 was higher than that obtained with the others dosimeters, possibly due to a smaller volume averaging effect. Ratio between the dose-rates determined with each dosimeter and those obtained with the film was evaluated for both treatment modalities. For the LGK, the ratio decreased as the dosimeter size increased and remained constant for collimator diameters larger than 8 mm. The same behaviour was observed for the linac and the ratio increased with field size, independent of the dosimeter used. These behaviours could be explained as an averaging volume effect due to dose gradient and lack of electronic equilibrium. Evaluation of the output factors for the LGK collimators indicated that, even when agreement was observed between Monte Carlo simulation and measurements with different dosimeters, this does not warrant that the absorbed dose to water rate in the field was properly known and thus, investigation of the reference dosimetry should be an important issue. These results indicated that alanine dosimeter provides a high degree of accuracy but cannot be used in fields smaller than 20 mm diameter. Gafchromic film can be considered as a suitable methodology for reference dosimetry. TLD dosimeters are not appropriate in fields

  3. Out of field dose during Gamma Knife treatment: a paediatric case study

    NASA Astrophysics Data System (ADS)

    Moutrie, V.; Grace, M.; Izard, M. A.; Fuller, J. W.

    2017-01-01

    An 11-year-old girl with an arteriovenous malformation (AVM) was referred for Gamma Knife treatment. As this would be the first paediatric treatment in Australia, additional investigations were undertaken into out of field dose to assure the best possible long term outcome for the patient. A phantom was constructed from water equivalent materials to simulate the patient. A target volume was defined to emulate the size and location of the AVM visible in diagnostic images. An ionisation chamber and EBT3 Gafchromic film were used to record absorbed dose at strategic points both on the surface and at depth within the phantom. On the day of treatment, EBT3 Gafchromic film was used to conduct in vivo dosimetry. The pre-treatment phantom measurements matched the planning system for the cranial section (the only modelled section) and no measurable dose above background was detected in the extracranial sites. In vivo measurements of the lenses returned doses of up to 2 cGy for imaging and 8 cGy for treatment which was also consistent with the planned dose. Dose to the thyroid, chest and abdomen was not measurable above background.

  4. Poster — Thur Eve — 39: Feasibility of Commissioning HybridArc with the Delta 4 two plane diode phantom: comparisons with Gafchromic Film

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Bojechko, C.; Ploquin, N.; University of Calgary, Department of Oncology, Tom Baker Cancer Center, Calgary AB

    2014-08-15

    HybridArc is a relatively novel radiation therapy technique which combines optimized dynamic conformai arcs (DCA) and intensity modulated radiation therapy (IMRT). HybridArc has possible dosimetry and efficiency advantages over stand alone DCA and IMRT treatments and can be readily implemented on any linac capable of DCA and IMRT, giving strong motivation to commission the modality. The Delta4 phantom (Scandidos, Uppsala, Sweden) has been used for IMRT and VMAT clinical dosimetric verification making it a candidate for HybridArc commissioning. However the HybridArc modality makes use of several non co-planar arcs which creates setup issues due to the geometry of the Delta4,more » resulting in possible phantom gantry collisions for plans with non-zero couch angles. An analysis was done determining the feasibility of using the Delta4 fixed at 0° couch angle compared with results obtained using Gafchromic ETB2 film (Ashland, Covington Kentucky) in an anthropomorphic phantom at the planned couch angles. A gamma index analysis of the measured and planned dose distributions was done using Delta4 and DoseLab Pro (Mobius Medical Systems, Houston Texas) software. For both arc and IMRT sub-fields there is reasonable correlation between the gamma index found from the Delta4 and Gafchromic film. All results show the feasibility of using the Delta4 for HybridArc commissioning.« less

  5. A comparison of TPS and different measurement techniques in small-field electron beams

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Donmez Kesen, Nazmiye, E-mail: nazo94@gmail.com; Cakir, Aydin; Okutan, Murat

    In recent years, small-field electron beams have been used for the treatment of superficial lesions, which requires small circular fields. However, when using very small electron fields, some significant dosimetric problems may occur. In this study, dose distributions and outputs of circular fields with dimensions of 5 cm and smaller, for nominal energies of 6, 9, and 15 MeV from the Siemens ONCOR Linac, were measured and compared with data from a treatment planning system using the pencil-beam algorithm in electron beam calculations. All dose distribution measurements were performed using the Gafchromic EBT film; these measurements were compared with datamore » that were obtained from the Computerized Medical Systems (CMS) XiO treatment planning system (TPS), using the gamma-index method in the PTW VeriSoft software program. Output measurements were performed using the Gafchromic EBT film, an Advanced Markus ion chamber, and thermoluminescent dosimetry (TLD). Although the pencil-beam algorithm is used to model electron beams in many clinics, there is no substantial amount of detailed information in the literature about its use. As the field size decreased, the point of maximum dose moved closer to the surface. Output factors were consistent; differences from the values obtained from the TPS were, at maximum, 42% for 6 and 15 MeV and 32% for 9 MeV. When the dose distributions from the TPS were compared with the measurements from the Gafchromic EBT films, it was observed that the results were consistent for 2-cm diameter and larger fields, but the outputs for fields of 1-cm diameter and smaller were not consistent. In CMS XiO TPS, calculated using the pencil-beam algorithm, the dose distributions of electron treatment fields that were created with circular cutout of a 1-cm diameter were not appropriate for patient treatment and the pencil-beam algorithm is not convenient for monitor unit (MU) calculations in electron dosimetry.« less

  6. A comparison of TPS and different measurement techniques in small-field electron beams.

    PubMed

    Donmez Kesen, Nazmiye; Cakir, Aydin; Okutan, Murat; Bilge, Hatice

    2015-01-01

    In recent years, small-field electron beams have been used for the treatment of superficial lesions, which requires small circular fields. However, when using very small electron fields, some significant dosimetric problems may occur. In this study, dose distributions and outputs of circular fields with dimensions of 5cm and smaller, for nominal energies of 6, 9, and 15MeV from the Siemens ONCOR Linac, were measured and compared with data from a treatment planning system using the pencil-beam algorithm in electron beam calculations. All dose distribution measurements were performed using the Gafchromic EBT film; these measurements were compared with data that were obtained from the Computerized Medical Systems (CMS) XiO treatment planning system (TPS), using the gamma-index method in the PTW VeriSoft software program. Output measurements were performed using the Gafchromic EBT film, an Advanced Markus ion chamber, and thermoluminescent dosimetry (TLD). Although the pencil-beam algorithm is used to model electron beams in many clinics, there is no substantial amount of detailed information in the literature about its use. As the field size decreased, the point of maximum dose moved closer to the surface. Output factors were consistent; differences from the values obtained from the TPS were, at maximum, 42% for 6 and 15MeV and 32% for 9MeV. When the dose distributions from the TPS were compared with the measurements from the Gafchromic EBT films, it was observed that the results were consistent for 2-cm diameter and larger fields, but the outputs for fields of 1-cm diameter and smaller were not consistent. In CMS XiO TPS, calculated using the pencil-beam algorithm, the dose distributions of electron treatment fields that were created with circular cutout of a 1-cm diameter were not appropriate for patient treatment and the pencil-beam algorithm is not convenient for monitor unit (MU) calculations in electron dosimetry. Copyright © 2015 American Association of

  7. EBT-XD Radiochromic Film Sensitivity Calibrations Using Proton Beams from a Pelletron Accelerator

    NASA Astrophysics Data System (ADS)

    Stockler, Barak; Grun, Alexander; Brown, Gunnar; Klein, Matthew; Wood, Jacob; Cooper, Anthony; Ward, Ryan; Freeman, Charlie; Padalino, Stephen; Regan, S. P.; Sangster, T. C.

    2017-10-01

    Radiochromic film (RCF) is a transparent detector film that permanently changes color following exposure to ionizing radiation. RCF is used frequently in medical applications, but also has been used in a variety of high energy density physics diagnostics. RCF is convenient to use because it requires no chemical processing and can be scanned using commercially available document scanners. In this study, the sensitivity of Gafchromic™ EBT-XD RCF to protons and x-rays was measured. Proton beams produced by the SUNY Geneseo Pelletron accelerator were directed into an evacuated target chamber where they scattered off a thin gold foil. The scattered protons were incident on a sample of RCF which subtended a range of angles around the scattering center. A new analysis method, which relies on the variation in scattered proton fluence as a function of scattering angle in accordance with the Rutherford scattering law, is currently being developed to speed up the proton calibrations. Samples of RCF were also exposed to x-ray radiation using an X-RAD 160 x-ray irradiator, allowing the sensitivity of RCF to X-rays to be measured. This work was funded in part by a Grant from the DOE through the Laboratory for Laser Energetics as well as the NSF.

  8. Comparative Analysis of Different Measurement Techniques for MLC Characterization: Preliminary Results

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Larraga-Gutierrez, J. M.; Ballesteros-Zebadua, P.; Garcia-Garduno, O. A.

    2008-08-11

    Radiation transmission, leakage and beam penumbra are essential dosimetric parameters related to the commissioning of a multileaf collimation system. This work shows a comparative analysis of commonly used film detectors: X-OMAT V2 and EDR2 radiographic films, and GafChromic EBT registered radiochromic film. The results show that X-OMAT over-estimates radiation leakage and 80-20% beam penumbra. However, according to the reference values reported by the manufacturer for these dosimetric parameters, all three films are adequate for MLC dosimetric characterization, but special care must be taken when X-OMAT V2 film is used due to its low energy photon dependence.

  9. Fractionated stereotactic radiotherapy: a method to evaluate geometric and dosimetric uncertainties using radiochromic films.

    PubMed

    Coscia, Gianluca; Vaccara, Elena; Corvisiero, Roberta; Cavazzani, Paolo; Ruggieri, Filippo Grillo; Taccini, Gianni

    2009-07-01

    In the authors' hospital, stereotactic radiotherapy treatments are performed with a Varian Clinac 600C equipped with a BrainLAB m3 micro-multileaf-collimator generally using the dynamic conformal arc technique. Patient immobilization during the treatment is achieved with a fixation mask supplied by BrainLAB, made with two reinforced thermoplastic sheets fitting the patient's head. With this work the authors propose a method to evaluate treatment geometric accuracy and, consequently, to determine the amount of the margin to keep in the CTV-PTV expansion during the treatment planning. The reproducibility of the isocenter position was tested by simulating a complete treatment on the anthropomorphic phantom Alderson Rando, inserting in between two phantom slices a high sensitivity Gafchromic EBT film, properly prepared and calibrated, and repeating several treatment sessions, each time removing the fixing mask and replacing the film inside the phantom. The comparison between the dose distributions measured on films and computed by TPS, after a precise image registration procedure performed by a commercial piece of software (FILMQA, 3cognition LLC (Division of ISP), Wayne, NJ), allowed the authors to measure the repositioning errors, obtaining about 0.5 mm in case of central spherical PTV and about 1.5 mm in case of peripheral irregular PTV. Moreover, an evaluation of the errors in the registration procedure was performed, giving negligible values with respect to the quantities to be measured. The above intrinsic two-dimensional estimate of treatment accuracy has to be increased for the error in the third dimension, but the 2 mm margin the authors generally use for the CTV-PTV expansion seems adequate anyway. Using the same EBT films, a dosimetric verification of the treatment planning system was done. Measured dose values are larger or smaller than the nominal ones depending on geometric irradiation conditions, but, in the authors' experimental conditions, always

  10. Radiotherapy dose verification on a customised head and neck perspex phantom

    NASA Astrophysics Data System (ADS)

    Eng, K. Y.; Kandaiya, S.; Yahaya, N. Z.

    2017-05-01

    IMRT dose planned for head and neck radiotherapy was verified using a customised acrylic head-and-neck phantom. The dosimeters used were calibrated Gafchromic EBT2 film and metal-oxide-semiconductor-field-effect-transistor (MOSFET). Target volumes (TV) and organs-at-risk (OAR) which were previously contoured by an oncologist on selected nasopharynx (NPC) patients were transferred to this phantom by an image fusion procedure. Three radiotherapy plans were done: Plan1 with 7-fields intensity-modulated radiotherapy (IMRT) of prescribed dose 70 Gy using 33 fractions; Plan2 with 7-fields IMRT plan at 70 Gy and 35 fractions; and Plan3 which was a mid-plane-dose (MPD) plan of 66 Gy at 33 fractions. The dose maps were first verified using MapCheck2 by SNC-PatientTM software. The passing rates from gamma analysis were 97.7% (Plan1), 93.1% (Plan2) and 100% (Plan3). Percentage difference between Treatment Planning System (TPS) calculated dose and MOSFET measured dose was comparatively higher than those from EBT2. Calculated dose and EBT2 measured doses showed differences of within the range of ±3% for TV and <±10% for OARs. However MOSFET had differences of within the range of ±6% for TV and within the range of ±10% for OARs between measured and planned doses. An overdose treatment may occur as TPS calculated doses were lower than the measured doses in these plans. This may be due to the effects of leaf leakage, leaf scatter and photon backscatter into the measuring tools (Pawlicki et al., 1999 and Ma et al., 2000). More IMRT plans have to be studied to validate this conclusion. However, the dose measurements were still within the 10% tolerance (AAPM Task Group 119). In conclusion, both GafchromicEBT2 film and MOSFET are suitable for IMRT radiotherapy dosimetry.

  11. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wilcox, Ellen; Daskalov, George; Nedialkova, Lucy

    Intensity-modulated radiotherapy (IMRT) treatment plan verification is often done using Kodak EDR2 film and a Vidar Dosimetry PRO trade mark sign film digitizer. However, since many hospitals are moving towards a filmless environment, access to a film processor may not be available. Therefore, we have investigated a newly available Gafchromic[reg] EBT film for IMRT dosimetry. Planar IMRT dose distributions are delivered to both EBT and EDR2 film and scanned with the Vidar VXR-16 as well as an Epson Expression 1680 flatbed scanner. The measured dose distributions are then compared to those calculated with a Pinnacle treatment planning system. The IMRTmore » treatments consisted of 7-9 6 MV beams for treatment of prostate, head and neck, and a few other sites. The films were analyzed using FilmQA trade mark sign (3cognition LLC) software. Comparisons between measured and calculated dose distributions are reported as dose difference (DD) (pixels within {+-}5%), distance to agreement (DTA) (3 mm), as well as gamma values ({gamma}) (dose={+-}3%, dist.=2 mm). Using EDR2 with the Vidar scanner is an established technique and agreement between calculated and measured dose distributions was better than 90% in all indices (DD, DTA, and {gamma}). However, agreement with calculations deteriorated reaching the lower 80% for EBT film scans with the Vidar scanner in logarithmic mode. The EBT Vidar scans obtained in linear mode showed an improved agreement to the upper 80% range, but artifacts were still observed across the scan. These artifacts were very distinct in all EBT scans and can be attributed to the way the film is transported through the scanner. In the Epson scanner both films are rigidly immobilized and the light source scans over the film. It was found that the Epson scanner performed equally well with both types of film giving agreement to better than 90% in all indices.« less

  12. "Characterization of ELEKTA SRS cone collimator using high spatial resolution monolithic silicon detector array".

    PubMed

    Shukaili, Khalsa Al; Corde, Stéphanie; Petasecca, Marco; Pereveratylo, Vladimir; Lerch, Michael; Jackson, Michael; Rosenfeld, Anatoly

    2018-05-22

    To investigate the accuracy of the dosimetry of radiation fields produced by small ELEKTA cone collimators used for stereotactic radiosurgery treatments (SRS) using commercially available detectors EBT3 Gafchromic TM film, IBA Stereotactic diode (SFD), and the recently developed detector DUO, which is a monolithic silicon orthogonal linear diode array detector. These three detectors were used for the measurement of beam profiles, output factors, and percentage depth dose for SRS cone collimators with cone sizes ranging from 5 to 50 mm diameter. The measurements were performed at 10 cm depth and 90 cm SSD. The SRS cone beam profiles measured with DUO, EBT3 film, and IBA SFD agreed well, results being in agreement within ±0.5 mm in the FWHM, and ±0.7 mm in the penumbra region. The output factor measured by DUO with 0.5 mm air gap above agrees within ±1% with EBT3. The OF measured by IBA SFD (corrected for the over-response) agreed with both EBT3 and DUO within ±2%. All three detectors agree within ±2% for PDD measurements for all SRS cones. The characteristics of the ELEKTA SRS cone collimator have been evaluated by using a monolithic silicon high spatial resolution detector DUO, EBT3, and IBA SFD diode. The DUO detector is suitable for fast real-time quality assurance dosimetry in small radiation fields typical for SRS/SRT. This has been demonstrated by its good agreement of measured doses with EBT 3 films. © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  13. SU-E-T-485: In Vivo Dosimetry with EBT3 Radiochromic Films for TBI Treatments

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lozares, S; Gracia, M; Olasolo, J

    2015-06-15

    Purpose: Total body irradiation (TBI) is a technique that requires special equipment to control “in vivo” the dose to the patient because it is a complex technique performed in extraordinary conditions. There are several devices to perform this task (diodes, TLDs, ionization chambers, MOSFET). In this paper we study the possibility of performing these measurements with radiochromic films EBT3 properly calibrated. This method has been compared to the PTW diodes system for TBI. Methods: Once made the TC to the patients, we measured different thicknesses of the relevant areas of the body (head, neck, chest with or without arms, umbilicusmore » area, knees and ankles); for each of these thicknesses we measured dose rate (cGy / UM) in RW3 phantom, in TBI conditions, with ionization chamber in the center; in turn, the input diode and the output of each configuration is placed to assign dose to each set of diodes. Movie calibration is performed according to manufacturer’s recommendations but TBI conditions. The dose at the center of each thickness compared to a linear interpolation of the dose at the entrance and exit, resulting in an adequate approximation. Finally in each session for each patient put a piece of film (2×2 cm2) at the entrance and another at the exit in each area, obtaining these readings and interpolating the estimated center dose, as with the diodes. Results: These results show a greater homogeneity in the distribution for use with film and validate the use of the same for this task and, if necessary, to avoid purchasing diode group if they have not. Conclusion: By using radiochromic films for this technique gives us a proper calculation of the dose received by the patient in the absence of other methods, or gives us a second additional track that already used normally.« less

  14. Developing new extension of GafChromic RTQA2 film to patient quality assurance field using a plan-based calibration method

    NASA Astrophysics Data System (ADS)

    Peng, Jiayuan; Zhang, Zhen; Wang, Jiazhou; Xie, Jiang; Chen, Junchao; Hu, Weigang

    2015-10-01

    GafChromic RTQA2 film is a type of radiochromic film designed for light field and radiation field alignment. The aim of this study is to extend the application of RTQA2 film to the measurement of patient specific quality assurance (QA) fields as a 2D relative dosimeter. Pre-irradiated and post-irradiated RTQA2 films were scanned in reflection mode using a flatbed scanner. A plan-based calibration (PBC) method utilized the mapping information of the calculated dose image and film grayscale image to create a dose versus pixel value calibration model. This model was used to calibrate the film grayscale image to the film relative dose image. The dose agreement between calculated and film dose images were analyzed by gamma analysis. To evaluate the feasibility of this method, eight clinically approved RapidArc cases (one abdomen cancer and seven head-and-neck cancer patients) were tested using this method. Moreover, three MLC gap errors and two MLC transmission errors were introduced to eight Rapidarc cases respectively to test the robustness of this method. The PBC method could overcome the film lot and post-exposure time variations of RTQA2 film to get a good 2D relative dose calibration result. The mean gamma passing rate of eight patients was 97.90%  ±  1.7%, which showed good dose consistency between calculated and film dose images. In the error test, the PBC method could over-calibrate the film, which means some dose error in the film would be falsely corrected to keep the dose in film consistent with the dose in the calculated dose image. This would then lead to a false negative result in the gamma analysis. In these cases, the derivative curve of the dose calibration curve would be non-monotonic which would expose the dose abnormality. By using the PBC method, we extended the application of more economical RTQA2 film to patient specific QA. The robustness of the PBC method has been improved by analyzing the monotonicity of the derivative of the

  15. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fontainha, C.C.P.; Nolasco, A.V.; Santos, A.P.

    Radiation dosimetry is commonly used to prevent deterministic radiation effects in high dose medical procedures. Radiochromic films find nowadays widely application in radiotherapy, interventional procedures and CT exams for isodose and maximum skin dose measurements. Moreover the size of the irradiated area and its distribution can be performed through the reading of the individual components in the RGB-spectrum. Particularly, radiochromic film has multiple advantages over alternative dosimeters for low-kV X-rays dosimetry. Concerned to spatial resolution it is far superior to that of ionization chambers and thermoluminescent dosimeters. For high energy photon fields (keV to MeV) the most used radiochromic filmmore » commercially available belongs to the EBT Gafchromic{sup R} series. On the other hand, for low energy photon fields in the x-ray range (20 kVp to 200 kVp) the best choice belongs to the XR-QA Gafchromic{sup R} film series. In this work we demonstrate the possibility of generating 2D images of thin polymeric composites films using EBT3 and XR-QA2 Gafchromic{sup R} films exposed to 6 MeV and 40 keV x-ray photons, respectively, using the digital filtering tools of the ImageJ{sup R} free software. In this context, EBT3 films were placed on the surface of a rigid anthropomorphic phantom. Then, they were covered with a thermoplastic mask made of PCL polymer. This setup was then exposed to 2.0 Gy absorbed dose in the Linear Accelerator beam. The EBT3 films were then scanned in the high resolution mode in a commercial scanner and the images subsequently treated with digital filters. It is somehow possible to see the image of the thermoplastic mask in the scanned image. However, in the treated image it is easy to observe the mask arrangement. The unexpected phenomenon here is the EB3 film ability to detect the attenuation of high energy photons by a plastic material, which in turn has a very low mass-energy attenuation coefficient, producing a very clear 2D

  16. SU-F-T-173: One-Scan Protocol: Verifying the Delivery of Spot-Scanning Proton Beam

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chan, M; Li, J; Chen, C

    Purpose: Radiochromic film for spot-scanning QA provides high spatial resolution and efficiency gains from one-shot irradiation for multiple depths. However, calibration can be a tedious procedure which may limit widespread use. Moreover, since there may be an energy dependence, which manifests as a depth dependence, this may require additional measurements for each patient. We present a one-scan protocol to simplify the procedure. Methods: We performed the calibration using an EBT3 film at depths of 18, 20, 24cm of Plastic Water exposed by a 6-level step-wedge plan on a Proteus Plus proton system (IBA, Belgium). The calibration doses ranged 65–250 cGy(RBE)more » for proton energies of 170–200MeV. A clinical prostate+nodes plan was used for validation. The planar doses at selected depths were measured with EBT3 films and analyzed using one-scan protocol (one-scan digitization of QA film and at least one film exposed to known dose). The Gamma passing rates, dose-difference maps, and profiles of 2D planar doses measured with EBT3 film, IBA MatriXX PT, versus TPS calculations were analyzed and compared. Results: The EBT3 film measurement results matched well with the TPS calculation data with an average passing rate of ∼95% for 2%/2mm and slightly lower passing rates were obtained from an ion chamber array detector. We were able to demonstrate that the use of a proton step-wedge provided clinically acceptable results and minimized variations between film-scanner orientation, inter-scan, and scanning conditions. Furthermore, it could be derived from no more than two films exposed to known doses (one could be zero) for rescaling the master calibration curve at each depth. Conclusion: The use of a proton step-wedge for calibration of EBT3 film increases efficiency. The sensitivity of the calibration to depth variations has been explored. One-scan protocol results appear to be comparable to that of the ion chamber array detector. One author has a research grant

  17. Feasibility of employing thick microbeams from superficial and orthovoltage kVp x-ray tubes for radiotherapy of superficial cancers

    NASA Astrophysics Data System (ADS)

    Kamali-Zonouzi, P.; Shutt, A.; Nisbet, A.; Bradley, D. A.

    2017-11-01

    Preclinical investigations of thick microbeams show these to be feasible for use in radiotherapeutic dose delivery. To create the beams we access a radiotherapy x-ray tube that is familiarly used within a conventional clinical environment, coupling this with beam-defining grids. Beam characterisation, both single and in the form of arrays, has been by use of both MCNP simulation and direct Gafchromic EBT film dosimetry. As a first step in defining optimal exit-beam profiles over a range of beam energies, simulation has been made of the x-ray tube and numbers of beam-defining parallel geometry grids, the latter being made to vary in thickness, slit separation and material composition. For a grid positioned after the treatment applicator, and of similar design to those used in the first part of the study, MCNP simulation and Gafchromic EBT film were then applied in examining the resultant radiation profiles. MCNP simulations and direct dosimetry both show useful thick microbeams to be produced from the x-ray tube, with peak-to-valley dose ratios (PVDRs) in the approximate range 8.8-13.9. Although the potential to create thick microbeams using radiotherapy x-ray tubes and a grid has been demonstrated, Microbeam Radiation Therapy (MRT) would still need to be approved outside of the preclinical setting, a viable treatment technique of clinical interest needing to benefit for instance from substantially improved x-ray tube dose rates.

  18. Development and characterization of a three-dimensional radiochromic film stack dosimeter for megavoltage photon beam dosimetry.

    PubMed

    McCaw, Travis J; Micka, John A; DeWerd, Larry A

    2014-05-01

    Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry. A film stack dosimeter was developed using Gafchromic(®) EBT2 films. The dosimeter consists of 22 films separated by 1 mm-thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed-dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage-depth-dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity-modulated radiation therapy (IMRT) procedure. The absorbed-dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under-response when the beam axis is parallel to the film

  19. Development and characterization of a three-dimensional radiochromic film stack dosimeter for megavoltage photon beam dosimetry

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    McCaw, Travis J., E-mail: mccaw@wisc.edu; Micka, John A.; DeWerd, Larry A.

    Purpose: Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry. Methods: A film stack dosimeter was developed using Gafchromic{sup ®} EBT2 films. The dosimeter consists of 22 films separated bymore » 1 mm-thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed-dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage-depth-dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity-modulated radiation therapy (IMRT) procedure. Results: The absorbed-dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under

  20. Extracorporeal blood treatment (EBT) methods in SIRS/Sepsis.

    PubMed

    Bellomo, R; Honoré, P M; Matson, J; Ronco, C; Winchester, J

    2005-05-01

    Extracorporeal blood purification treatment (EBT) methods have been used in the treatment of experimental and human SIRS/Sepsis in a variety of settings and with variable reports of efficacy and safety. Their role in the management of SIRS/Sepsis remains controversial. To develop consensus statements regarding important aspects of research, practice and technical management concerning EBT. Systematic review of published study. Evidence-based grading of information available. Consensus development regarding fundamental questions about EBT. Consensus was achieved on all questions posed during the conference. It was agreed that there is currently a clear biological rational for EBT in SIRS and Sepsis. It was agreed that conventional CVVH has sufficiently been shown not to be effective in SIRS/Sepsis in the absence of concomitant ARF and that other therapies such as plasmapheresis or high-volume hemofiltration or coupled plasma filtration adsorption appear more promising and should be tested in multicentre randomized controlled trials. Patients with ARF and SIRS/Sepsis should be treated differently from those with ARF alone even though current practice in this field is not well known. Patients with refractory septic shock should be considered for EBT. Appropriate end points for clinical trials can be defined and chosen according to the goals of the trial. Different technologies exist for EBT and better understanding of the merits and safety of each is needed as well as better standardization of methodology and dose. Broad consensus can be achieved on several aspects of EBT and can be used to inform clinical practice and to help guide the establishment of a future research agenda.

  1. Advanced radiochromic film methodologies for quantitative dosimetry of small and nonstandard fields

    NASA Astrophysics Data System (ADS)

    Rosen, Benjamin S.

    Radiotherapy treatments with small and nonstandard fields are increasing in use as collimation and targeting become more advanced, which spare normal tissues while increasing tumor dose. However, dosimetry of small and nonstandard fields is more difficult than that of conventional fields due to loss of lateral charged-particle equilibrium, tight measurement setup requirements, source occlusion, and the volume-averaging effect of conventional dosimeters. This work aims to create new small and nonstandard field dosimetry protocols using radiochromic film (RCF) in conjunction with novel readout and analysis methodologies. It also is the intent of this work to develop an improved understanding of RCF structure and mechanics for its quantitative use in general applications. Conventional digitization techniques employ white-light, flatbed document scanners or scanning-laser densitometers which are not optimized for RCF dosimetry. A point-by-point precision laser densitometry system (LDS) was developed for this work to overcome the film-scanning artifacts associated with the use of conventional digitizers, such as positional scan dependence, off-axis light scatter, glass bed interference, and low signal-to-noise ratios. The LDS was shown to be optically traceable to national standards and to provide highly reproducible density measurements. Use of the LDS resulted in increased agreement between RCF dose measurements and the single-hit detector model of film response, facilitating traceable RCF calibrations based on calibrated physical quantities. GafchromicRTM EBT3 energy response to a variety of reference x-ray and gamma-ray beam qualities was also investigated. Conventional Monte Carlo methods are not capable of predicting film intrinsic energy response to arbitrary particle spectra. Therefore, a microdosimetric model was developed to simulate the underlying physics of the radiochromic mechanism and was shown to correctly predict the intrinsic response relative to a

  2. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rodriguez, M; Bartolac, S; Rezaee, M

    Purpose: To examine the agreement between absorbed doses calculated by RayStation treatment planning algorithm to those measured with gafchromic film and ion chamber when the photon beam is perturbed by attenuation or lateral scatter of lung material. Methods: A gafchromic EBT2 film was placed in the center of a 30×30×20 cm{sup 3} solid water phantom with a 5 cm lung slab placed at 10 cm depth. The film was irradiated at SSD = 100 cm with a 6 MV photon beam, 10×10 and 5×5 cm{sup 2} field sizes and with the beam parallel to the film and lung slab. Amore » CT was performed to the phantom arrangement for RayStation dose calculation. The films were scanned in an Epson 10000X flatbed scanner and analyzed using the red channel, 16 bits, 76 dpi. PDD curves at the central axis and profiles at dmax were also measured in water using a CC13 (0.13 CC) ion chamber. Measurements and calculation of PDD curves at the central axis and profiles at dmax and 20 cm depth were compared using the criteria suggested by the AAPM Task Group # 53. Results: The PDD curves measured with gafchromic film and those measured in water with ion chamber agree with the ones calculated by RayStation within the uncertainty of the measurements which is within 3%. The passing rate values of the measured and calculated profiles for the 2 field sizes are within 94% for both at dmax and at 20 cm depth. Conclusion: Raystation dose calculation engine models inhomogeneity corrections. Differences between the calculated PDD curves and profiles and those measured with gafchromic film are within the uncertainty of the measurements and inside of the agreement tolerance suggested by TG53. Therefore, RayStation treatment planning has an acceptable algorithm to correct dose delivered by photon beams perturbed by lung tissue.« less

  3. Small Radiation Beam Dosimetry for Radiosurgery of Trigeminal Neuralgia: One Case Analysis

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Garcia-Garduno, O. A.; Larraga-Gutierrez, J. M.; Unidad de Radioneurocirugia, Instituto Nacional de Neurologia y Neurocirugia. Insurgentes Sur 3677, Col. La Fama, C. P. 14269, Tlalpan, Mexico, D. F.

    2008-08-11

    The use of small radiation beams for trigeminal neuralgia (TN) treatment requires high precision and accuracy in dose distribution calculations and delivery. Special attention must be kept on the type of detector to be used. In this work, the use of GafChromic EBT registered radiochromic and X-OMAT V2 radiographic films for small radiation beam characterization is reported. The dosimetric information provided by the films (total output factors, tissue maximum ratios and off axis ratios) is compared against measurements with a shielded solid state (diode) reference detector. The film dosimetry was used for dose distribution calculations for the treatment of trigeminalmore » neuralgia radiosurgery. Comparison of the isodose curves shows that the dosimetry produced with the X-OMAT radiographic film overestimates the dose distributions in the penumbra region.« less

  4. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Medina, L; Adrada, A; Filipuzzi, M

    Purpose: The purpose of this paper is to characterize EBT3 using two types of scanner, analyzing the factors of influence of each dosimetry system. Methods: The film used in this study was GAFCHROMIC EBT3, the films were exposed at a dose range between 0Gy a 9Gy in a solid water phantom, SSD=100cm, 5cm depth and perpendicularly to the 6MV photon beam generated by a Novalis TX linear accelerator equipped with an HDMLC. A Farmer type ion chamber TN30013 (PTW) was used to determine the dose delivered to the film. The films were digitized with a scanner EPSON expression 10000XL andmore » the VIDAR DosimetryPro Adventage RED. Software RIT113v6.1 was used for construction of the calibration curve and analysis. The film characteristics investigated were: response at different dose levels, sensitivity to orientation and side and resolution through the results of the spatial response function by analyzing a step pattern. Additionally, 20 IMRT treatment fields were measured with both scanner and compared with calculated dose using gamma index analysis (3%-3mm). Results: The OD obtained for dose level 2Gy in the orientation portrait of the film on the scanner EPSON is (0,222±0,19) and for Vidar RED (0,252±0,10) and landscape is for EPSON (0,211±0,25) and for Vidar RED (0,250±0,11) . The orientation dependence with respect to film side is about 0,09% for EPSON and about 0.03% for VIDAR. The spatial response function increase in response to the Gaussian function FWHM EPSON scanner (0.18mm) compared with VIDAR scanner function (less than 0.06mm) was observed. We analyzed 20 total plan dose distributions the number of pixels with gamma>1 (3%-3mm) was 0.7%±1.2 [0.1%; 2.82%] for EBT3-VIDAR y 2%±2.9 [0.2%; 3.5%] for EBT3-EPSON. Conclusion: VIDAR scanner shows better sensitivity. EBT3 film shows a different response between portrait and landscape orientation. Step pattern is better reproduce by VIDAR scanner.« less

  5. Poster – 13: Evaluation of an in-house CCD camera film dosimetry imaging system for small field deliveries

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lalonde, Michel; Alexander, Kevin; Olding, Tim

    Purpose: Radiochromic film dosimetry is a standard technique used in clinics to verify modern conformal radiation therapy delivery, and sometimes in research to validate other dosimeters. We are using film as a standard for comparison as we improve high-resolution three-dimensional gel systems for small field dosimetry; however, precise film dosimetry can be technically challenging. We report here measurements for fractionated stereotactic radiation therapy (FSRT) delivered using volumetric modulated arc therapy (VMAT) to investigate the accuracy and reproducibility of film measurements with a novel in-house readout system. We show that radiochromic film can accurately and reproducibly validate FSRT deliveries and alsomore » benchmark our gel dosimetry work. Methods: VMAT FSRT plans for metastases alone (PTV{sub MET}) and whole brain plus metastases (WB+PTV{sub MET}) were delivered onto a multi-configurational phantom with a sheet of EBT3 Gafchromic film inserted mid-plane. A dose of 400 cGy was prescribed to 4 small PTV{sub MET} structures in the phantom, while a WB structure was prescribed a dose of 200 cGy in the WB+PTV{sub MET} iterations. Doses generated from film readout with our in-house system were compared to treatment planned doses. Each delivery was repeated multiple times to assess reproducibility. Results and Conclusions: The reproducibility of film optical density readout was excellent throughout all experiments. Doses measured from the film agreed well with plans for the WB+PTV{sub MET} delivery. But, film doses for PTV{sub MET} only deliveries were significantly below planned doses. This discrepancy is due to stray/scattered light perturbations in our system during readout. Corrections schemes will be presented.« less

  6. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wen, N; Lu, S; Qin, Y

    Purpose: To evaluate the dosimetric uncertainty associated with Gafchromic (EBT3) films and establish an absolute dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). Methods: EBT3 films were irradiated at each of seven different dose levels between 1 and 15 Gy with open fields, and standard deviations of dose maps were calculated at each color channel for evaluation. A scanner non-uniform response correction map was built by registering and comparing film doses to the reference diode array-based dose map delivered with the same doses. To determine the temporal dependence of EBT3 films, the average correction factors of differentmore » dose levels as a function of time were evaluated up to four days after irradiation. An integrated film dosimetry protocol was developed for dose calibration, calibration curve fitting, dose mapping, and profile/gamma analysis. Patient specific quality assurance (PSQA) was performed for 93 SRS/SBRT treatment plans. Results: The scanner response varied within 1% for the field sizes less than 5 × 5 cm{sup 2}, and up to 5% for the field sizes of 10 × 10 cm{sup 2}. The scanner correction method was able to remove visually evident, irregular detector responses found for larger field sizes. The dose response of the film changed rapidly (∼10%) in the first two hours and plateaued afterwards, ∼3% change between 2 and 24 hours. The mean uncertainties (mean of the standard deviations) were <0.5% over the dose range 1∼15Gy for all color channels for the OD response curves. The percentage of points passing the 3%/1mm gamma criteria based on absolute dose analysis, averaged over all tests, was 95.0 ± 4.2. Conclusion: We have developed an absolute film dose dosimetry protocol using EBT3 films. The overall uncertainty has been established to be approximately 1% for SRS and SBRT PSQA. The work was supported by a Research Scholar Grant, RSG-15-137-01-CCE from the American Cancer Society.« less

  7. Development and demonstration of 2D dosimetry using optically stimulated luminescence from new Al2O3 films for radiotherapy applications

    NASA Astrophysics Data System (ADS)

    Ahmed, Md Foiez

    Scope and Method of Study: The goal of this work was to develop and demonstrate a 2D dosimetry system based on the optically stimulated luminescence (OSL) from new Al2O3 films for radiotherapy applications. A 2D laser-scanning system was developed for the readout and two OSL films (Al2O3:C and Al2O3:C,Mg) were tested. A dose reconstruction algorithm addressing corrections required for the characteristic material properties and the properties related to the system design was developed. The dosimetric properties of the system were tested using clinical X-ray (6 MV) beam. The feasibility of small field dosimetry was tested using heavy ion beams (221 MeV proton and 430 MeV 12C beam). For comparison, clinical tests were performed with ionization chamber, diode arrays and the commercial radiochromic films (Gafchromic EBT3) when applicable. Findings and Conclusions: The results demonstrate that the developed image reconstruction algorithm enabled > 300x faster laser-scanning readout of the Al2O3 films, eliminating the restriction imposed by its slow luminescence decay. The algorithm facilitates submillimeter spatial resolution, reduces the scanner position dependence (of light collection efficiency) and removes the inherent galvo geometric distortion, among other corrections. The system has a background signal < 1 mGy, linearity correction factor of < 10% up to ˜4.0 Gy and < 2% dose uncertainty over the clinically relevant dose range of 0.1 - 30 Gy. The system has a dynamic range of 4 - 5 orders, only limited by PMT linearity. The absolute response from Al2O2:C films is higher than Al2O 2:C,Mg films, but with lower image signal-to-noise ratio due to lower concentration of fast F+-center emission. As a result, Al2O2:C,Mg films are better suited than Al2O3:C films for small field dosimetry, which requires precise dosimetry with sub-millimeter spatial resolution. The dose uncertainty associated with OSL film dosimetry is lower than that associated with EBT3 film dosimetry

  8. MAGAT gel and EBT2 film‐based dosimetry for evaluating source plugging‐based treatment plan in Gamma Knife stereotactic radiosurgery

    PubMed Central

    Vivekanandhan, S.; Kale, S.S.; Rath, G.K.; Senthilkumaran, S.; Thulkar, S.; Subramani, V.; Laviraj, M.A.; Bisht, R.K.; Mahapatra, A.K.

    2012-01-01

    This work illustrates a procedure to assess the overall accuracy associated with Gamma Knife treatment planning using plugging. The main role of source plugging or blocking is to create dose falloff in the junction between a target and a critical structure. We report the use of MAGAT gel dosimeter for verification of an experimental treatment plan based on plugging. The polymer gel contained in a head‐sized glass container simulated all major aspects of the treatment process of Gamma Knife radiosurgery. The 3D dose distribution recorded in the gel dosimeter was read using a 1.5T MRI scanner. Scanning protocol was: CPMG pulse sequence with 8 equidistant echoes, TR=7 s, echo step=14 ms, pixel size=0.5 mm x 0.5 mm, and slice thickness of 2 mm. Using a calibration relationship between absorbed dose and spin‐spin relaxation rate (R2), we converted R2 images to dose images. Volumetric dose comparison between treatment planning system (TPS) and gel measurement was accomplished using an in‐house MATLAB‐based program. The isodose overlay of the measured and computed dose distribution on axial planes was in close agreement. Gamma index analysis of 3D data showed more than 94% voxel pass rate for different tolerance criteria of 3%/2 mm, 3%/1 mm and 2%/2 mm. Film dosimetry with GAFCHROMIC EBT 2 film was also performed to compare the results with the calculated TPS dose. Gamma index analysis of film measurement for the same tolerance criteria used for gel measurement evaluation showed more than 95% voxel pass rate. Verification of gamma plan calculated dose on account of shield is not part of acceptance testing of Leksell Gamma Knife (LGK). Through this study we accomplished a volumetric comparison of dose distributions measured with a polymer gel dosimeter and Leksell GammaPlan (LGP) calculations for plans using plugging. We propose gel dosimeter as a quality assurance (QA) tool for verification of plug‐based planning. PACS number: 87.53.Ly, 87.55.‐x, 87

  9. SU-F-T-493: An Investigation Into the Feasibility of Using PipsPro Software with Film for Linac QA

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Underwood, R

    2016-06-15

    Purpose: To determine the feasibility of using radiochromic and radiographic film with Pipspro software for quality assurance of linear accelerators with no on-board imaging. Methods: The linear accelerator being used is a Varian Clinac 21EX. All IGRT is performed using the BrainLab ExacTrac system. Because of the lack of on board imaging, certain monthly and annual TG-142 quality assurance tests are more difficult to perform and analyze to a high degree of accuracy. Pipspro was not designed to be used with hard film, and to our knowledge its use with film had not been investigated. The film used will bemore » GafChromic EBT3 film and Kodak EDR2 film, scanned with an Epson V700 scanner. The following routine tests will be attempted: MLC picket fence, light vs. radiation field coincidence, starshots, and MLC transmission. Results: The only tests that gave accurate and reliable results were the couch, gantry, and collimator starshots. Typical MV and kV images are acquired with a much higher level of contrast between the irradiated and non-irradiated areas when compared to film. Pipspro relies on this level of contrast to be able to automatically detect the fiducial points from its phantom devices, leaf edges for picket fence and transmission tests, and jaw edges for light vs. radiation field tests. Because of this, certain tests gave erroneous results and others were not able to be performed in the software at all, with either type of film. The number of monitor units delivered to the film, the experimental setup, and the scan settings was not able to rectify the problem. Conclusion: For linear accelerators with no on-board imaging, it is not recommended to use hard film with PipsPro to perform TG-142 quality assurance tests. Other software or methods should instead be investigated.« less

  10. Patient-Specific QA of Spot-Scanning Proton Beams using Radiochromic Film.

    PubMed

    Chan, Maria F; Chen, Chin-Cheng; Shi, Chengyu; Li, Jingdong; Tang, Xiaoli; Li, Xiang; Mah, Dennis

    2017-05-01

    Radiochromic film for spot-scanning QA provides high spatial resolution and efficiency gains from one-shot irradiation for multiple depths. However, calibration can be a tedious procedure which may limit widespread use. Moreover, since there may be an energy dependence, which manifests as a depth dependence, this may require additional measurements for each patient. We present a one-scan protocol to simplify the procedure. A calibration using an EBT3 film, exposed by a 6-level step-wedge plan on a Proteus ® PLUS proton system (IBA, Belgium), was performed at depths of 18, 20, 24cm using Plastic Water ® (CIRS, Norfolk, VA). The calibration doses ranged from 65-250 cGy(RBE) (relative biological effectiveness) for proton energies of 170-200 MeV. A clinical prostate+nodes plan was used for validation. The planar doses at selected depths were measured with EBT3 films and analyzed using One-scan protocol (one-scan digitization of QA film and at least one film exposed to a known dose). The gamma passing rates, dose-difference maps, and profiles of 2D planar doses measured with EBT3 film and IBA MatriXX-PT, versus the RayStation TPS calculations were analyzed and compared. The EBT3 film measurement results matched well with the TPS calculation data with an average passing rate of ~95% for 2%/2mm and slightly lower passing rates were obtained from an ion chamber array detector. We were able to demonstrate that the use of a proton step-wedge provided clinically acceptable results and minimized variations between film-scanner orientation, inter-scan, and scanning conditions. Furthermore, for relative dosimetry (calibration is not done at the time of experiment) it could be derived from no more than two films exposed to known doses (one could be zero) for rescaling the master calibration curve at each depth. The sensitivity of the calibration to depth variations has been explored. One-scan protocol results appear to be comparable to that of the ion chamber array detector

  11. Calibration of GafChromic XR-RV3 radiochromic film for skin dose measurement using standardized x-ray spectra and a commercial flatbed scanner.

    PubMed

    McCabe, Bradley P; Speidel, Michael A; Pike, Tina L; Van Lysel, Michael S

    2011-04-01

    In this study, newly formulated XR-RV3 GafChromic film was calibrated with National Institute of Standards and Technology (NIST) traceability for measurement of patient skin dose during fluoroscopically guided interventional procedures. The film was calibrated free-in-air to air kerma levels between 15 and 1100 cGy using four moderately filtered x-ray beam qualities (60, 80, 100, and 120 kVp). The calibration films were scanned with a commercial flatbed document scanner. Film reflective density-to-air kerma calibration curves were constructed for each beam quality, with both the orange and white sides facing the x-ray source. A method to correct for nonuniformity in scanner response (up to 25% depending on position) was developed to enable dose measurement with large films. The response of XR-RV3 film under patient backscattering conditions was examined using on-phantom film exposures and Monte Carlo simulations. The response of XR-RV3 film to a given air kerma depended on kVp and film orientation. For a 200 cGy air kerma exposure with the orange side of the film facing the source, the film response increased by 20% from 60 to 120 kVp. At 500 cGy, the increase was 12%. When 500 cGy exposures were performed with the white side facing the x-ray source, the film response increased by 4.0% (60 kVp) to 9.9% (120 kVp) compared to the orange-facing orientation. On-phantom film measurements and Monte Carlo simulations show that using a NIST-traceable free-in-air calibration curve to determine air kerma in the presence of backscatter results in an error from 2% up to 8% depending on beam quality. The combined uncertainty in the air kerma measurement from the calibration curves and scanner nonuniformity correction was +/- 7.1% (95% C.I.). The film showed notable stability. Calibrations of film and scanner separated by 1 yr differed by 1.0%. XR-RV3 radiochromic film response to a given air kerma shows dependence on beam quality and film orientation. The presence of

  12. Evaluation and implementation of triple‐channel radiochromic film dosimetry in brachytherapy

    PubMed Central

    Bradley, David; Nisbet, Andrew

    2014-01-01

    like effects and to provide reliable dosimetric results. We have demonstrated that radiochromic film dosimetry with GAFCHROMIC EBT3 film and a commercial flatbed scanner is a viable method for brachytherapy dose distribution measurement, and uncertainties may be reduced with triple‐channel dosimetry and specific film scan and evaluation methodologies. PACS numbers: 87.55.Qr, 87.56.bg, 87.55.km PMID:25207417

  13. [A practical procedure to improve the accuracy of radiochromic film dosimetry: a integration with a correction method of uniformity correction and a red/blue correction method].

    PubMed

    Uehara, Ryuzo; Tachibana, Hidenobu; Ito, Yasushi; Yoshino, Shinichi; Matsubayashi, Fumiyasu; Sato, Tomoharu

    2013-06-01

    It has been reported that the light scattering could worsen the accuracy of dose distribution measurement using a radiochromic film. The purpose of this study was to investigate the accuracy of two different films, EDR2 and EBT2, as film dosimetry tools. The effectiveness of a correction method for the non-uniformity caused from EBT2 film and the light scattering was also evaluated. In addition the efficacy of this correction method integrated with the red/blue correction method was assessed. EDR2 and EBT2 films were read using a flatbed charge-coupled device scanner (EPSON 10000G). Dose differences on the axis perpendicular to the scanner lamp movement axis were within 1% with EDR2, but exceeded 3% (Maximum: +8%) with EBT2. The non-uniformity correction method, after a single film exposure, was applied to the readout of the films. A corrected dose distribution data was subsequently created. The correction method showed more than 10%-better pass ratios in dose difference evaluation than when the correction method was not applied. The red/blue correction method resulted in 5%-improvement compared with the standard procedure that employed red color only. The correction method with EBT2 proved to be able to rapidly correct non-uniformity, and has potential for routine clinical IMRT dose verification if the accuracy of EBT2 is required to be similar to that of EDR2. The use of red/blue correction method may improve the accuracy, but we recommend we should use the red/blue correction method carefully and understand the characteristics of EBT2 for red color only and the red/blue correction method.

  14. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Richter, Christian; Pawelke, Joerg; Karsch, Leonhard

    Purpose: The aim of this article is to investigate the energy dependence of the radiochromic film type, Gafchromic EBT-1, when scanned with a flatbed scanner for film readout. Methods: Dose response curves were determined for 12 different beam qualities ranging from a 10 kVp x-ray beam to a 15 MVp x-ray beam and include also two high energy electron beam qualities (6 and 18 MeV). The dose responses measured as net optical density (netOD) for the different beam qualities were normalized to the response of a reference beam quality (6 MVp). Results: A strong systematic energy dependence of the filmmore » response was found. The lower the effective beam energy, the less sensitive the EBT-1 films get. The maximum decrease in dose for the same film response between the 25 kVp and 6 MVp beam qualities was 44%. Additionally, a difference in energy dependence for different doses was discovered, meaning that higher doses show a smaller dependency on energy than lower doses. The maximum decrease in the normalized netOD was found to be 25% for a dose of 0.5 Gy relative to the normalized netOD for 10 Gy. Moreover, a scaling procedure is introduced, allowing the correction of the energy dependence for the investigated beam qualities and also for comparable x-ray beam qualities within the energy range studied. Conclusions: A strong energy dependence for EBT-1 radiochromic films was found. The films were readout with a flatbed scanner. If the effective beam energy is known, the energy dependence can be corrected with the introduced scaling procedure. Further investigation of the influence of the spectral band of the readout device on energy dependence is needed to understand the reason for the different energy dependences found in this and previous works.« less

  15. Deriving detector-specific correction factors for rectangular small fields using a scintillator detector.

    PubMed

    Qin, Yujiao; Zhong, Hualiang; Wen, Ning; Snyder, Karen; Huang, Yimei; Chetty, Indrin J

    2016-11-08

    The goal of this study was to investigate small field output factors (OFs) for flat-tening filter-free (FFF) beams on a dedicated stereotactic linear accelerator-based system. From this data, the collimator exchange effect was quantified, and detector-specific correction factors were generated. Output factors for 16 jaw-collimated small fields (from 0.5 to 2 cm) were measured using five different detectors including an ion chamber (CC01), a stereotactic field diode (SFD), a diode detector (Edge), Gafchromic film (EBT3), and a plastic scintillator detector (PSD, W1). Chamber, diodes, and PSD measurements were performed in a Wellhofer water tank, while films were irradiated in solid water at 100 cm source-to-surface distance and 10 cm depth. The collimator exchange effect was quantified for rectangular fields. Monte Carlo (MC) simulations of the measured configurations were also performed using the EGSnrc/DOSXYZnrc code. Output factors measured by the PSD and verified against film and MC calculations were chosen as the benchmark measurements. Compared with plastic scintillator detector (PSD), the small volume ion chamber (CC01) underestimated output factors by an average of -1.0% ± 4.9% (max. = -11.7% for 0.5 × 0.5 cm2 square field). The stereotactic diode (SFD) overestimated output factors by 2.5% ± 0.4% (max. = 3.3% for 0.5 × 1 cm2 rectangular field). The other diode detector (Edge) also overestimated the OFs by an average of 4.2% ± 0.9% (max. = 6.0% for 1 × 1 cm2 square field). Gafchromic film (EBT3) measure-ments and MC calculations agreed with the scintillator detector measurements within 0.6% ± 1.8% and 1.2% ± 1.5%, respectively. Across all the X and Y jaw combinations, the average collimator exchange effect was computed: 1.4% ± 1.1% (CC01), 5.8% ± 5.4% (SFD), 5.1% ± 4.8% (Edge diode), 3.5% ± 5.0% (Monte Carlo), 3.8% ± 4.7% (film), and 5.5% ± 5.1% (PSD). Small field detectors should be used with caution with a clear understanding of their

  16. SU-F-T-08: Brachytherapy Film Dosimetry in a Water Phantom for a Ring and Tandem HDR Applicator

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lee, B; Grelewicz, Z; Kang, Z

    2016-06-15

    Purpose: The feasibility of dose measurement using new generation EBT3 film was explored in a water phantom for a ring and tandem HDR applicator for measurements tracking mucosal dose during cervical brachytherapy. Methods: An experimental fixture was assembled to position the applicator in a water phantom. Prior to measurement, calibration curves for EBT3 film in water and in solidwater were verified. EBT3 film was placed at different known locations around the applicator in the water tank. A CT scan of the phantom with applicator was performed using clinical protocol. A typical cervical cancer treatment plan was then generated by Oncentramore » brachytherapy planning system. A dose of 500 cGy was prescribed to point A (2 cm, 2 cm). Locations measured by film included the outer surface of the ring, measurement point A-m (2.2 cm, 2.2 cm), and profiles extending from point A-m parallel to the tandem. Three independent measurements were conducted. The doses recorded by film were carefully analyzed and compared with values calculated by the treatment planning system. Results: Assessment of the EBT3 films indicate that the dose at point A matches the values predicted by the planning system. Dose to the point A-m was 411.5 cGy, and the outer circumferential surface dose of the ring was between 500 and 1150 cGy. It was found that from the point A-m, the dose drops 60% within 4.5 cm on the line parallel to the tandem. The measurement doses agree with the treatment planning system. Conclusion: Use of EBT3 film is feasible for in-water measurements for brachytherapy. A carefully machined apparatus will likely improve measurement accuracy. In a typical plan, our study found that the ring surface dose can be 2.5 times larger than the point A prescription dose. EBT3 film can be used to monitor mucosal dose in brachytherapy treatments.« less

  17. Implementation of small field radiotherapy dosimetry for spinal metastase case

    NASA Astrophysics Data System (ADS)

    Rofikoh, Wibowo, W. E.; Pawiro, S. A.

    2017-07-01

    The main objective of this study was to know dose profile of small field radiotherapy in the spinal metastase case with source axis distance (SAD) techniques. In addition, we evaluated and compared the dose planning of stereotactic body radiation therapy (SBRT) and conventional techniques to measurements with Exradin A16 and Gafchromic EBT3 film dosimeters. The results showed that film EBT3 had a highest precision and accuracy with the average of the standard deviation of ±1.7 and maximum discrepancy of 2.6 %. In addition, the average value of Full Wave Half Maximum (FWHM) and its largest deviation in small field size of 0.8 x 0.8 cm2 are 0.82 cm and 16.3 % respectively, while it was found around 2.36 cm and 3 % for the field size of 2.4 x 2.4 cm2. The comparison between penumbra width and the collimation was around of 37.1 % for the field size of 0.8 x 0.8 cm2, while it was found of 12.4 % for the field size of 2.4 x 2.4 cm2.

  18. Evaluation of material heterogeneity dosimetric effects using radiochromic film for COMS eye plaques loaded with {sup 125}I seeds (model I25.S16)

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Acar, Hilal; Chiu-Tsao, Sou-Tung; Oezbay, Ismail

    Purpose: (1) To measure absolute dose distributions in eye phantom for COMS eye plaques with {sup 125}I seeds (model I25.S16) using radiochromic EBT film dosimetry. (2) To determine the dose correction function for calculations involving the TG-43 formalism to account for the presence of the COMS eye plaque using Monte Carlo (MC) method specific to this seed model. (3) To test the heterogeneous dose calculation accuracy of the new version of Plaque Simulator (v5.3.9) against the EBT film data for this seed model. Methods: Using EBT film, absolute doses were measured for {sup 125}I seeds (model I25.S16) in COMS eyemore » plaques (1) along the plaque's central axis for (a) uniformly loaded plaques (14-20 mm in diameter) and (b) a 20 mm plaque with single seed, and (2) in off-axis direction at depths of 5 and 12 mm for all four plaque sizes. The EBT film calibration was performed at {sup 125}I photon energy. MC calculations using MCNP5 code for a single seed at the center of a 20 mm plaque in homogeneous water and polystyrene medium were performed. The heterogeneity dose correction function was determined from the MC calculations. These function values at various depths were entered into PS software (v5.3.9) to calculate the heterogeneous dose distributions for the uniformly loaded plaques (of all four sizes). The dose distributions with homogeneous water assumptions were also calculated using PS for comparison. The EBT film measured absolute dose rate values (film) were compared with those calculated using PS with homogeneous assumption (PS Homo) and heterogeneity correction (PS Hetero). The values of dose ratio (film/PS Homo) and (film/PS Hetero) were obtained. Results: The central axis depth dose rate values for a single seed in 20 mm plaque measured using EBT film and calculated with MCNP5 code (both in ploystyrene phantom) were compared, and agreement within 9% was found. The dose ratio (film/PS Homo) values were substantially lower than unity (mostly between 0

  19. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hadsell, M; Holcombe, C; Chin, E

    Introduction: As diagnostic techniques become more sensitive and targeting methods grow in accuracy, target volumes continue to shrink and SBRT becomes more prevalent. Due to this fact, patient-specific QA must also enhance resolution and accuracy in order to verify dose delivery in these volumes. It has been suggested that when measuring small fields at least two separate detectors be used to verify delivered dose. Therefore, we have instituted a secondary patient QA verification for small (<3cm) SBRT fields using Gafchromic EBT2 film. Methods: Films were cross-calibrated using a Farmer chamber in plastic water at reference conditions as defined by TG-51.more » Films were scanned, and an RGB calibration curve was created according to best practices published by Ashland, Inc. Four SBRT cases were evaluated both with the Scandidos Delta4 and with EBT2 films sandwiched in plastic water. Raw values obtained from the film were converted to dose using an in-house algorithm employing all three color channels to increase accuracy and dosimetric range. Gamma and dose profile comparisons to Eclipse dose calculations were obtained using RIT and compared to values obtained with the Delta4. Results: Film gamma pass rates at 2% and 2mm were similar to those obtained with the Delta4. However, dose difference histograms showed better absolute dose agreement, with the average mean film dose agreeing with calculation to 0.3% and the Delta4 only agreeing to 3.1% across the cases. Additionally, films provided more resolution than the Delta4 and thus their dose profiles better succeeded in diagnosing dose calculation inaccuracies. Conclusion: We believe that the implementation of secondary patient QA using EBT2 film analyzed with all three color channels is an invaluable tool for evaluation of small SBRT fields. Furthermore, we have shown that this method can sometimes provide a more detailed and faithful reproduction of plan dose than the Delta4.« less

  20. SU-F-T-568: QA of a Multi-Target Multi-Dose VMAT SRS

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Roa, D; Kuo, J; Gonzales, A

    2016-06-15

    Purpose: To, experimentally, corroborated the prescribed doses utilizing dosimeters (e.g. films and TLDs) that can provide high spatial resolution, allow dose measurement of multiple targets at once, and provide accurate dosimetric results. Methods: A single-isocenter 6FFF SRS VMAT plan consisting of one 358° arc at 0° couch angle and four 179° arcs at 30°, 60°, 330° and 300° couch angles respectively, was generated in ECLIPSE v.11 using a Rando-Alderson anthropomorphic head phantom CT study. This plan was a reproduction of a clinical plan generated for a stage-IV melanoma patient diagnosed with 19 intracranial lesions. The phantom was loaded with axiallymore » mounted (between phantom slabs) Gafchromic EBT3 film and TLDs strategically positioned within various target volumes. Film and TLDS were calibrated according to established protocols. Target prescription doses were 16 Gy (3cc≤, 3 lesions), 18 Gy (∼1–3cc, 10 lesions) and 20 Gy (≤1cc, 6 lesions). Phantom setup was verified through CBCT imaging prior to irradiation. Gafchromic films were scanned in transmission mode and TLDs were read, respectively, ∼24 hrs after irradiation. Results: Dose calibrated Gafchromic film data were compared to the ECLIPSE calculated data using a 3% / 3mm gamma function analysis. Results for the gamma values were 96–99% in agreement with the calculated data and with 84–90% of the film pixels within the 3% dose difference. TLD data showed a dose difference of 0.4–8% while the film data for those same locations yielded a difference of 0.4–4%. It was observed that the highest dose discrepancies correlated with the location of the small volume targets. Conclusion: Overall this study corroborated that a VMAT SRS treatment, employing various treatment table rotations and arcs, to multiple intracranial lesions with multiple dose prescriptions can be delivered accurately with the existing radiotherapy technology.« less

  1. SU-F-T-578: Characterization of Vidar DosimetryPro Advantage RED Scanner with Application to SBRT and SRS QA

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Dumas, M; Wen, N

    Purpose: To use Gafchromic EBT3 film to quantify key dosimetric characteristics of the Vidar DosimetryPro Advantage RED film scanner for use in SBRT/SRS QA, by analyzing scanner uniformity and dose sensitivity. Method: Gafchromic EBT3 film was used in this study. Films were irradiated using 6MV FFF and 10MV FFF beams from a Varian Edge linear accelerator, with setup of 100cm SAD at depth 5 cm. Nine doses were delivered per film, with calibration dose ranges of 1–10 Gy and 3–24 Gy for 6MV FFF, and 3–27 Gy for 10MV FFF. Films were scanned with the long side of the filmmore » parallel to the detector array. Dose calibration curves were fitted to a 3rd degree polynomial. The derivative of a calibration curve was taken to determine the scanner’s sensitivity per dose delivered (OD/Gy). Scanner non-uniformity was calculated in 2 dimensions by taking the mean of standard deviation in each row and column. Absolute dose SRS/SBRT Gamma analyses were performed with passing criteria of 3% and 1mm DTA. For comparison, Gamma analyses were also performed using an Epson Expression 10000 XL. Results: Uniformity for the Vidar scanner was 0.37% +/− 0.03% in the perpendicular to scan direction and 0.67% +/− 0.05% in the parallel to scan direction, with an overall uniformity of 0.52% +/− 0.03%. Epson red channel uniformity was 0.85% +/− 0.05% and 0.88% +/− 0.08% for the green channel. The Vidar average dose sensitivity from 1–10 Gy was 0.112 OD/Gy and 0.061 OD/Gy for 3–24 Gy. SBRT/SRS Gamma pass rates were 97.8 +/− 1.4 for Vidar and 97.5 +/− 1.4 for Epson. Conclusion: The Vidar scanner has 41% less non-uniformity compared to Epson XL10000 green channel. The dose sensitivity is 2–3 time greater for the Vidar scanner compared to the Epson in the SRS/SBRT dose range of 5–24 Gy.« less

  2. The artefacts of radiochromic film dosimetry with flatbed scanners and their causation by light scattering from radiation-induced polymers.

    PubMed

    Schoenfeld, Andreas A; Poppinga, Daniela; Harder, Dietrich; Doerner, Karl-Joachim; Poppe, Bjoern

    2014-07-07

    Optical experiments and theoretical considerations have been undertaken in order to understand the causes of the 'orientation effect' and the 'parabola effect', the artefacts impairing the desired light absorption measurement on radiochromic EBT3 films with flatbed scanners. EBT3 films exposed to doses up to 20.9 Gy were scanned with an Epson Expression 10000XL flatbed scanner in landscape and portrait orientation. The horizontally and vertically polarized light components of the scanner were determined, and another Epson Expression 10000XL flatbed scanner was disassembled to examine its optical components. The optical properties of exposed and unexposed EBT3 films were studied with incident polarized and unpolarized white light, and the transmitted red light was investigated for its polarization and scattering properties including the distribution of the scattering angles. Neutral density filters were studied for comparison. Guidance was sought from the theory of light scattering from rod-like macromolecular structures. The drastic dose-dependent variation of the transmitted total light current as function of the orientation of front and rear polarizers, interpreted by light scattering theory, shows that the radiation-induced polymerization of the monomers of EBT3 films produces light scattering oscillators preferably polarized at right angles with the coating direction of the film. The directional distribution of the scattered light is partly anisotropic, with a preferred scattering plane at right angles with the coating direction, indicating light scattering from stacks of coherently vibrating oscillators piled up along the monomer crystals. The polyester carrier film also participates in these effects. The 'orientation' and 'parabola' artefacts due to flatbed scanning of radiochromic films can be explained by the interaction of the polarization-dependent and anisotropic light scattering from exposed and unexposed EBT3 films with the quantitative difference

  3. Fabrication of prototype for measuring the exhaled breath temperature (EBT) to support detection of asthma

    NASA Astrophysics Data System (ADS)

    Harnawan, A. A.; Mariati; Fahrudin, A.; Assegaf, A.

    2017-05-01

    Recently, EBT has been proven as the marker of airways inflammation like asthma and proposed as the non-invasive tool. Although EBT device has already been made but this device is rarely used for patients in Indonesia. The aim of this study is to develop a prototype which accurately measures EBT and is comfortably used by patients including children. This prototype was made using SHT11 as a sensor of EBT which is integrated on a thermal flask 0.5L. This flask filled up patient breath and the temperature of air breath was measured. The EBT of twelve healthy samples and seven samples with asthma was examined using this prototype, the measurement was done within three minutes for all of them. The test results of EBT on healthy samples obtained the median is 33.9°C within of 33.0°C - 34.7°C and EBT on asthma samples obtained median is 35.0°C within the range 34.9°C - 36.0°C.

  4. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Gagnon, Jean-Christophe; Theriault, Dany; Guillot, Mathieu

    Purpose: To compare the performance of plastic scintillation detectors (PSD) for quality assurance (QA) in stereotactic radiosurgery conditions to a microion-chamber (IC), Gafchromic EBT2 films, 60 008 shielded photon diode (SD) and unshielded diodes (UD), and assess a new 2D crosshair array prototype adapted to small field dosimetry. Methods: The PSD consists of a 1 mm diameter by 1 mm long scintillating fiber (BCF-60, Saint-Gobain, Inc.) coupled to a polymethyl-methacrylate optical fiber (Eska premier, Mitsubishi Rayon Co., Ltd., Tokyo, Japan). Output factors (S{sub c,p}) for apertures used in radiosurgery ranging from 4 to 40 mm in diameter have been measured.more » The PSD crosshair array (PSDCA) is a water equivalent device made up of 49 PSDs contained in a 1.63 cm radius area. Dose profiles measurements were taken for radiosurgery fields using the PSDCA and were compared to other dosimeters. Moreover, a typical stereotactic radiosurgery treatment using four noncoplanar arcs was delivered on a spherical phantom in which UD, IC, or PSD was placed. Using the Xknife planning system (Integra Radionics Burlington, MA), 15 Gy was prescribed at the isocenter, where each detector was positioned. Results: Output Factors measured by the PSD have a mean difference of 1.3% with Gafchromic EBT2 when normalized to a 10 x 10 cm{sup 2} field, and 1.0% when compared with UD measurements normalized to the 35 mm diameter cone. Dose profiles taken with the PSD crosshair array agreed with other single detectors dose profiles in spite of the presence of the 49 PSDs. Gamma values comparing 1D dose profiles obtained with PSD crosshair array with Gafchromic EBT2 and UD measured profiles shows 98.3% and 100.0%, respectively, of detector passing the gamma acceptance criteria of 0.3 mm and 2%. The dose measured by the PSD for a complete stereotactic radiosurgery treatment is comparable to the planned dose corrected for its SD-based S{sub c,p} within 1.4% and 0.7% for 5 and 35 mm

  5. SU-E-T-665: Radiochromic Film Quenching Effect Reduction for Proton Beam Dosimetry

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Aldelaijan, S; Alzorkany, F; Moftah, B

    2015-06-15

    Purpose: Depending on the useful dose range in which radiochromic films operate, number of different radiochromic film models have been designed. The impact of different film models on quenching effect for percent depth dose (PDD) measurements in proton beams has been investigated. Methods: Calibrated PTW Markus ionization chamber was used to measure PDD and beam output for 26.5 MeV protons produced by CS30 cyclotron. An aluminum cylinder was added in front of the beam exit serving as a radiation shutter. The measured signal was normalized to a monitor chamber reading and subsequently scaled by ratio of water-to-air stopping powers atmore » given depth, while the effective depth of measurements was scaled by ratios of material-to-water physical densities and CSDA ranges. Output was measured in water at 2.1 mm reference-depth in the plateau upstream from the Bragg peak. Following the TRS-398 reference dosimetry protocol for proton beams, the output was calibrated in water. Three radiochromic film models (EBT, EBT3 and HD-V2) were calibrated within Lexan phantom positioned at the same water-equivalent depth. Thicknesses of films sensitive layers were 34 µm, 30 µm and 8 µm, respectively. Small film pieces (1 x 2 cm{sup 2}) were positioned within polyethylene phantom along the beam central axis with an angulation of 5° for PDD measurements. Results: While the output of the proton beam was found to be around 7 Gy/sec, the actual value of the output per monitor chamber reading (2.32 Gy/nC) was used for reference-dose irradiations during film calibration. Dose ratios at the Bragg peak relative to the reference-depth were 3.88, 2.52, 2.19, and 2.02 for the Markus chamber, HD-V2, EBT3, and EBT film models, respectively. Conclusion: Results at hand suggest that quenching effect is reduced when a radiochromic film model with smaller sensitive layer thickness is used for PDD measurements in proton beams. David Lewis is the owner of RCF Consulting, LLC.« less

  6. SU-F-T-82: Dosimetric Evaluation of a Shield Used for Hemi-Body Skin Electron Irradiation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rivers, C; Singh, A; AlDahlawi, I

    Purpose: We had several mycosis fungoides patients with a limited disease to about half of the skin surface. A custom-made plywood shield was used to protect the non-targeted skin region with our total skin electron irradiation (TSEI) technique. We report a dosimetric evaluation for our “hemi-body” skin electron irradiation technique. Methods: The technique is similar to our clinical total skin electron irradiation (TSEI), performed with a six-pair dual field (Stanford technique) at an extended source-to-skin distance (SSD) of 377 cm, with the addition of a plywood shield placed 50 cm from the patient. The shield is made of three layersmore » of standard 5/8″ thick plywood (total thickness of 4.75 cm) that are clamped securely on an adjustable-height stand. Gafchromic EBT3 films were used in assessing the shield’s transmission factor and the extend of the dose penumbra region. To verify the dose delivered for hemi-body skin radiation in a real patient treatment, in-vivo dosimetry using Gafchromic EBT3 films were performed. Film pieces were taped on the patient skin to measure the dose received during the first two fractions, placed on the forehead and upper body (shielded region); and also at the level of pelvic area, left thigh, and left ankle. Results: The shield transmission factor was found to be 10%, and the width of the penumbra (80-to-20% dose fall-off) was about 12 cm. In-vivo dosimetry of a real case confirmed the expected shielded area dose. Conclusion: Hemi-Body skin electron irradiation at an extended SSD is feasible with the addition of a plywood shield at a distance from patient skin. The penumbra dose region and the shield’s transmission factor should be evaluated prior to clinical use. We have treated several hemi-body skin patients with our custom-made plywood shield, the current patient measurements are representative of these for other patients as well.« less

  7. In vivo dosimetry with radiochromic films in low-voltage intraoperative radiotherapy of the breast.

    PubMed

    Avanzo, M; Rink, A; Dassie, A; Massarut, S; Roncadin, M; Borsatti, E; Capra, E

    2012-05-01

    EBT2 radiochromic films were studied and used for in vivo dosimetry in targeted intraoperative radiotherapy (TARGIT), a technique in which the Intrabeam system (Carl Zeiss, Oberkochen, Germany) is used to perform intraoperative partial breast irradiation with x-rays of 50 kV(p). The energy of the radiation emitted by the Intrabeam with the different spherical applicators, under 1 and 2 cm of solid water, and under the tungsten impregnated rubber used for shielding of the heart in TARGIT of the breast, was characterized with measurements of half-value layer (HVL). The stability of response of EBT2 was verified inside this range of energies. EBT2 films were calibrated using the red and green channels of the absorption spectrum in the 0-20 Gy dose range delivered by the Intrabeam x-rays. The dependence of film response on temperature during irradiation was measured. For in vivo dosimetry, pieces of radiochromic films wrapped in sterile envelopes were inserted after breast conserving surgery and before TARGIT into the excision cavity, on the skin and on the shielded pectoralis fascia for treatments of the left breast. HVLs of the Intrabeam in TARGIT of the breast correspond to effective energies of 20.7-36.3 keV. The response of EBT2 was constant inside this range of energies. We measured the dose to the target tissue and to organs at risk in 23 patients and obtained an average dose of 13.52 ± 1.21 Gy to the target tissue. Dose to the skin in close proximity to the applicator was 2.22 ± 0.97 Gy, 0.29 ± 0.17 Gy at 5-10 cm from the applicator, and 0.08 ± 0.07 Gy at more than 10 cm from the applicator. Dose to the pectoral muscle for left breast treatment was 0.57 ± 0.23 Gy. Our results show that EBT2 films are accurate at the beam energies, dose range, and irradiation temperature found in TARGIT and that in vivo dosimetry in TARGIT with EBT2 films wrapped in sterile envelopes is a feasible procedure. Measured dose to the organs at risk indicates

  8. In vivo dosimetry with radiochromic films in low-voltage intraoperative radiotherapy of the breast

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Avanzo, M.; Rink, A.; Dassie, A.

    Purpose: EBT2 radiochromic films were studied and used for in vivo dosimetry in targeted intraoperative radiotherapy (TARGIT), a technique in which the Intrabeam system (Carl Zeiss, Oberkochen, Germany) is used to perform intraoperative partial breast irradiation with x-rays of 50 kV{sub p}. Methods: The energy of the radiation emitted by the Intrabeam with the different spherical applicators, under 1 and 2 cm of solid water, and under the tungsten impregnated rubber used for shielding of the heart in TARGIT of the breast, was characterized with measurements of half-value layer (HVL). The stability of response of EBT2 was verified inside thismore » range of energies. EBT2 films were calibrated using the red and green channels of the absorption spectrum in the 0-20 Gy dose range delivered by the Intrabeam x-rays. The dependence of film response on temperature during irradiation was measured. For in vivo dosimetry, pieces of radiochromic films wrapped in sterile envelopes were inserted after breast conserving surgery and before TARGIT into the excision cavity, on the skin and on the shielded pectoralis fascia for treatments of the left breast. Results: HVLs of the Intrabeam in TARGIT of the breast correspond to effective energies of 20.7-36.3 keV. The response of EBT2 was constant inside this range of energies. We measured the dose to the target tissue and to organs at risk in 23 patients and obtained an average dose of 13.52 {+-} 1.21 Gy to the target tissue. Dose to the skin in close proximity to the applicator was 2.22 {+-} 0.97 Gy, 0.29 {+-} 0.17 Gy at 5-10 cm from the applicator, and 0.08 {+-} 0.07 Gy at more than 10 cm from the applicator. Dose to the pectoral muscle for left breast treatment was 0.57 {+-} 0.23 Gy. Conclusions: Our results show that EBT2 films are accurate at the beam energies, dose range, and irradiation temperature found in TARGIT and that in vivo dosimetry in TARGIT with EBT2 films wrapped in sterile envelopes is a feasible procedure

  9. A quantitative method to the analysis of MLC leaf position and speed based on EPID and EBT3 film for dynamic IMRT treatment with different types of MLC.

    PubMed

    Li, Yinghui; Chen, Lixin; Zhu, Jinhan; Wang, Bin; Liu, Xiaowei

    2017-07-01

    A quantitative method based on the electronic portal imaging system (EPID) and film was developed for MLC position and speed testing; this method was used for three MLC types (Millennium, MLCi, and Agility MLC). To determine the leaf position, a picket fence designed by the dynamic (DMLC) model was used. The full-width half-maximum (FWHM) values of each gap measured by EPID and EBT3 were converted to the gap width using the FWHM versus nominal gap width relationship. The algorithm developed for the picket fence analysis was able to quantify the gap width, the distance between gaps, and each individual leaf position. To determine the leaf speed, a 0.5 × 20 cm 2 MLC-defined sliding gap was applied across a 14 × 20 cm 2 symmetry field. The linacs ran at a fixed-dose rate. The use of different monitor units (MUs) for this test led to different leaf speeds. The effect of leaf transmission was considered in a speed accuracy analysis. The difference between the EPID and film results for the MLC position is less than 0.1 mm. For the three MLC types, twice the standard deviation (2 SD) is provided; 0.2, 0.4, and 0.4 mm for gap widths of three MLC types, and 0.1, 0.2, and 0.2 mm for distances between gaps. The individual leaf positions deviate from the preset positions within 0.1 mm. The variations in the speed profiles for the EPID and EBT3 results are consistent, but the EPID results are slightly better than the film results. Different speeds were measured for each MLC type. For all three MLC types, speed errors increase with increasing speed. The analysis speeds deviate from the preset speeds within approximately 0.01 cm s -1 . This quantitative analysis of MLC position and speed provides an intuitive evaluation for MLC quality assurance (QA). © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  10. SU-F-T-562: Validation of EPID-Based Dosimetry for FSRS Commissioning

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Song, Y; Saleh, Z; Obcemea, C

    Purpose: The prevailing approach to frameless SRS (fSRS) small field dosimetry is Gafchromic film. Though providing continuous information, its intrinsic uncertainties in fabrication, response, scan, and calibration often make film dosimetry subject to different interpretations. In this study, we explored the feasibility of using EPID portal dosimetry as a viable alternative to film for small field dosimetry. Methods: Plans prescribed a dose of 21 Gy were created on a flat solid water phantom with Eclipse V11 and iPlan for small static square fields (1.0 to 3.0 cm). In addition, two clinical test plans were computed by employing iPlan on amore » CIRS Kesler head phantom for target dimensions of 1.2cm and 2.0cm. Corresponding portal dosimetry plans were computed using the Eclipse TPS and delivered on a Varian TrueBeam machine. EBT-XD film dosimetry was performed as a reference. The isocenter doses were measured using EPID, OSLD, stereotactic diode, and CC01 ion chamber. Results: EPID doses at the center of the square field were higher than Eclipse TPS predicted portal doses, with the mean difference being 2.42±0.65%. Doses measured by EBT-XD film, OSLD, stereotactic diode, and CC01 ion chamber revealed smaller differences (except OSLDs), with mean differences being 0.36±3.11%, 4.12±4.13%, 1.7±2.76%, 1.45±2.37% for Eclipse and −1.36±0.85%, 2.38±4.2%, −0.03±0.50%, −0.27±0.78% for iPlan. The profiles measured by EPID and EBT-XD film resembled TPS (Eclipse and iPlan) predicted ones within 3.0%. For the two clinical test plans, the EPID mean doses at the center of field were 2.66±0.68% and 2.33±0.32% higher than TPS predicted doses. Conclusion: We found that results obtained with EPID portal dosimetry were slightly higher (∼2%) than those obtained with EBT-XD film, diode, and CC01 ion chamber with the exception of OSLDs, but well within IROC tolerance (5.0%). Therefore, EPID has the potential to become a viable real-time alternative method to film

  11. Cell irradiation setup and dosimetry for radiobiological studies at ELBE

    NASA Astrophysics Data System (ADS)

    Zeil, K.; Beyreuther, E.; Lessmann, E.; Wagner, W.; Pawelke, J.

    2009-07-01

    The radiation source ELBE delivers different types of secondary radiation, which is used for cell irradiation studies in radiobiological research. Thereby an important issue is the determination of the biological effectiveness of photon radiation as a function of photon energy by using low-energetic, monochromatic channeling radiation (10-100 keV) and high-energetic bremsstrahlung (up to 40 MV). Radiobiological studies at the research facility ELBE demand special technical and dosimetric prerequisites. Therefore, a cell irradiation system (CIS) has been designed, constructed and installed at the beam line. The CIS allows automatic irradiation of a larger cell sample number and the compensation of spatial inhomogeneity of the dose distribution within the beam spot. The recently introduced GafChromic ® EBT radiochromic film model has been used to verify the cell irradiation dose deposition achieving a dose uncertainty of <5%. Both, the installed cell irradiation system and the developed dosimetric procedure based on the use of the EBT film have been experimentally tested at ELBE. The biological effectiveness of 34 MV bremsstrahlung with respect to 200 kV X-rays from a conventional X-ray tube has been determined. An RBE value of 0.75 has been measured in good agreement with literature.

  12. A reusable OSL-film for 2D radiotherapy dosimetry.

    PubMed

    Wouter, Crijns; Dirk, Vandenbroucke; Paul, Leblans; Tom, Depuydt

    2017-10-19

    Optical stimulated luminescence (OSL) combines reusability, sub-mm resolution, and a linear dose response in a single radiation detection technology. Such a combination is currently lacking in radiotherapy dosimetry. But OSL-films have a strong energy dependent response to keV photons due to a relative high effective atomic number (Z eff ). The current work studied the applicability of a 2D OSL-film with a reduced Z eff as (IMRT/VMAT) dosimeter. Based on their commercial OSL-film experience, Agfa Healthcare N.V. produced a new experimental OSL-film for RT dosimetry. This film had a lower effective atomic number compared to the films used in radiology. Typical 2D dosimeter requirements such as uniformity, dose response, signal stability with time, and angular dependence were evaluated. Additionally, the impact of a possible residual energy dependence was assessed for the infield as well as the out-of-field region of both static beams and standard intensity modulated patterns (chair and pyramid). The OSL-film's reusable nature allowed for a film specific absolute and linear calibration including a flood-field uniformity correction. The OSL-film was scanned with a CR-15X engine based reader using a strict timing (i.e. 4 min after 'beam on' or as soon as possible) to account for spontaneous recombination. The OSL-film had good basic response properties: non-uniformities  ⩽2.6%, a linear dose response (0-32 Gy), a linear signal decay (0.5% min -1 ) over the 20 min measured, and limited angular dependence  ⩽2.6%. Due to variations of the energy spectrum, larger dose differences were noted outside the central region of the homogenous phantom and outside both static and IMRT fields. However, the OSL-film's measured dose differences of the IMRT patterns were lower than those of Gafchromic EBT measurements ([-1.6%, 2.1%] versus [-2.9%, 3.6%]). The current OSL-film could be used as a reusable high resolution dosimeter with read-out immediately after irradiation

  13. Measurements of proton energy spectra using a radiochromic film stack

    NASA Astrophysics Data System (ADS)

    Filkins, T. M.; Steidle, Jessica; Ellison, D. M.; Steidle, Jeffrey; Freeman, C. G.; Padalino, S. J.; Fiksel, G.; Regan, S. P.; Sangster, T. C.

    2014-10-01

    The energy spectrum of protons accelerated from the rear-side of a thin foil illuminated with ultra-intense laser light from the OMEGA EP laser system at the University of Rochester's Laboratory for Laser Energetics (LLE) was measured using a stack of radiochromic film (RCF). The film stack consisted of four layers of Gafchromic HD-V2 film and four layers of Gafchromic MD-V2-55 film. Aluminum foils of various thicknesses were placed between each piece of RCF in the stack. This arrangement allowed protons with energies of 30 MeV to reach the back layer of RCF in the stack. The stack was placed in the detector plane of a Thomson parabola ion energy (TPIE) spectrometer. Each piece of film in the stack was scanned using a commercially available flat-bed scanner (Epson 10000XL). The resulting optical density was converted into proton fluence using an absolute calibration of the RCF obtained at the SUNY Geneseo 1.7 MV Pelletron accelerator laboratory. In these calibration measurements, the sensitivity of the radiochromic film was measured using monoenergetic protons produced by the accelerator. Details of the analysis procedure and the resulting proton energy spectra will be presented. Funded in part by a grant from the DOE through the Laboratory for Laser Energetics.

  14. Monte Carlo uncertainty analysis of dose estimates in radiochromic film dosimetry with single-channel and multichannel algorithms.

    PubMed

    Vera-Sánchez, Juan Antonio; Ruiz-Morales, Carmen; González-López, Antonio

    2018-03-01

    To provide a multi-stage model to calculate uncertainty in radiochromic film dosimetry with Monte-Carlo techniques. This new approach is applied to single-channel and multichannel algorithms. Two lots of Gafchromic EBT3 are exposed in two different Varian linacs. They are read with an EPSON V800 flatbed scanner. The Monte-Carlo techniques in uncertainty analysis provide a numerical representation of the probability density functions of the output magnitudes. From this numerical representation, traditional parameters of uncertainty analysis as the standard deviations and bias are calculated. Moreover, these numerical representations are used to investigate the shape of the probability density functions of the output magnitudes. Also, another calibration film is read in four EPSON scanners (two V800 and two 10000XL) and the uncertainty analysis is carried out with the four images. The dose estimates of single-channel and multichannel algorithms show a Gaussian behavior and low bias. The multichannel algorithms lead to less uncertainty in the final dose estimates when the EPSON V800 is employed as reading device. In the case of the EPSON 10000XL, the single-channel algorithms provide less uncertainty in the dose estimates for doses higher than four Gy. A multi-stage model has been presented. With the aid of this model and the use of the Monte-Carlo techniques, the uncertainty of dose estimates for single-channel and multichannel algorithms are estimated. The application of the model together with Monte-Carlo techniques leads to a complete characterization of the uncertainties in radiochromic film dosimetry. Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  15. Plan delivery quality assurance for CyberKnife: Statistical process control analysis of 350 film-based patient-specific QAs.

    PubMed

    Bellec, J; Delaby, N; Jouyaux, F; Perdrieux, M; Bouvier, J; Sorel, S; Henry, O; Lafond, C

    2017-07-01

    Robotic radiosurgery requires plan delivery quality assurance (DQA) but there has never been a published comprehensive analysis of a patient-specific DQA process in a clinic. We proposed to evaluate 350 consecutive film-based patient-specific DQAs using statistical process control. We evaluated the performance of the process to propose achievable tolerance criteria for DQA validation and we sought to identify suboptimal DQA using control charts. DQAs were performed on a CyberKnife-M6 using Gafchromic-EBT3 films. The signal-to-dose conversion was performed using a multichannel-correction and a scanning protocol that combined measurement and calibration in a single scan. The DQA analysis comprised a gamma-index analysis at 3%/1.5mm and a separate evaluation of spatial and dosimetric accuracy of the plan delivery. Each parameter was plotted on a control chart and control limits were calculated. A capability index (Cpm) was calculated to evaluate the ability of the process to produce results within specifications. The analysis of capability showed that a gamma pass rate of 85% at 3%/1.5mm was highly achievable as acceptance criteria for DQA validation using a film-based protocol (Cpm>1.33). 3.4% of DQA were outside a control limit of 88% for gamma pass-rate. The analysis of the out-of-control DQA helped identify a dosimetric error in our institute for a specific treatment type. We have defined initial tolerance criteria for DQA validations. We have shown that the implementation of a film-based patient-specific DQA protocol with the use of control charts is an effective method to improve patient treatment safety on CyberKnife. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  16. The origin of the flatbed scanner artifacts in radiochromic film dosimetry—key experiments and theoretical descriptions

    NASA Astrophysics Data System (ADS)

    Schoenfeld, Andreas A.; Wieker, Soeren; Harder, Dietrich; Poppe, Bjoern

    2016-11-01

    The optical origin of the lateral response and orientation artifacts, which occur when using EBT3 and EBT-XD radiochromic films together with flatbed scanners, has been reinvestigated by experimental and theoretical means. The common feature of these artifacts is the well-known parabolic increase in the optical density OD(x)  =  -log10 I(x)/I 0(x) versus offset x from the scanner midline (Poppinga et al 2014 Med. Phys. 41 021707). This holds for landscape and portrait orientations as well as for the three color channels. Dose-independent optical subjects, such as neutral density filters, linear polarizers, the EBT polyester foil and diffusive glass, also present the parabolic lateral artifact when scanned with a flatbed scanner. The curvature parameter c of the parabola function OD(x)  =  c 0  +  cx 2 is found to be a linear function of the dose, the parameters of which are influenced by the film orientation and film type, EBT3 or EBT-XD. The ubiquitous parabolic shape of function OD(x) is attributed (a) to the optical path-length effect (van Battum et al 2016 Phys. Med. Biol. 61 625-49), due to the increasing obliquity of the optical scanner light associated with increasing offset x from the scanner midline, and (b) and (c) to the partial polarization and scattering of the light leaving the film, which affect the ratio ~I(x)/{{I}0}(x) , thus making OD(x) increase with x 2. The orientation effect results from the changes of effects (b) and (c) associated with turning the film position, and thereby the orientation of the polymer structure of the sensitive film layer. In a comparison of experimental results obtained with selected optical subjects, the relative weights of the contributions of the optical path-length effect and the polarization and scattering of light leaving the films to the lateral response artifact have been estimated to be of the same order of magnitude. Mathematical models of these causes for the parabolic shape of function

  17. The origin of the flatbed scanner artifacts in radiochromic film dosimetry-key experiments and theoretical descriptions.

    PubMed

    Schoenfeld, Andreas A; Wieker, Soeren; Harder, Dietrich; Poppe, Bjoern

    2016-11-07

    The optical origin of the lateral response and orientation artifacts, which occur when using EBT3 and EBT-XD radiochromic films together with flatbed scanners, has been reinvestigated by experimental and theoretical means. The common feature of these artifacts is the well-known parabolic increase in the optical density OD(x)  =  -log 10 I(x)/I 0 (x) versus offset x from the scanner midline (Poppinga et al 2014 Med. Phys. 41 021707). This holds for landscape and portrait orientations as well as for the three color channels. Dose-independent optical subjects, such as neutral density filters, linear polarizers, the EBT polyester foil and diffusive glass, also present the parabolic lateral artifact when scanned with a flatbed scanner. The curvature parameter c of the parabola function OD(x)  =  c 0   +  cx 2 is found to be a linear function of the dose, the parameters of which are influenced by the film orientation and film type, EBT3 or EBT-XD. The ubiquitous parabolic shape of function OD(x) is attributed (a) to the optical path-length effect (van Battum et al 2016 Phys. Med. Biol. 61 625-49), due to the increasing obliquity of the optical scanner light associated with increasing offset x from the scanner midline, and (b) and (c) to the partial polarization and scattering of the light leaving the film, which affect the ratio [Formula: see text], thus making OD(x) increase with x 2 . The orientation effect results from the changes of effects (b) and (c) associated with turning the film position, and thereby the orientation of the polymer structure of the sensitive film layer. In a comparison of experimental results obtained with selected optical subjects, the relative weights of the contributions of the optical path-length effect and the polarization and scattering of light leaving the films to the lateral response artifact have been estimated to be of the same order of magnitude. Mathematical models of these causes for the parabolic shape of

  18. A reusable OSL-film for 2D radiotherapy dosimetry

    NASA Astrophysics Data System (ADS)

    Wouter, Crijns; Dirk, Vandenbroucke; Paul, Leblans; Tom, Depuydt

    2017-11-01

    Optical stimulated luminescence (OSL) combines reusability, sub-mm resolution, and a linear dose response in a single radiation detection technology. Such a combination is currently lacking in radiotherapy dosimetry. But OSL-films have a strong energy dependent response to keV photons due to a relative high effective atomic number (Z eff). The current work studied the applicability of a 2D OSL-film with a reduced Z eff as (IMRT/VMAT) dosimeter. Based on their commercial OSL-film experience, Agfa Healthcare N.V. produced a new experimental OSL-film for RT dosimetry. This film had a lower effective atomic number compared to the films used in radiology. Typical 2D dosimeter requirements such as uniformity, dose response, signal stability with time, and angular dependence were evaluated. Additionally, the impact of a possible residual energy dependence was assessed for the infield as well as the out-of-field region of both static beams and standard intensity modulated patterns (chair and pyramid). The OSL-film’s reusable nature allowed for a film specific absolute and linear calibration including a flood-field uniformity correction. The OSL-film was scanned with a CR-15X engine based reader using a strict timing (i.e. 4 min after ‘beam on’ or as soon as possible) to account for spontaneous recombination. The OSL-film had good basic response properties: non-uniformities  ⩽2.6%, a linear dose response (0-32 Gy), a linear signal decay (0.5% min-1) over the 20 min measured, and limited angular dependence  ⩽2.6%. Due to variations of the energy spectrum, larger dose differences were noted outside the central region of the homogenous phantom and outside both static and IMRT fields. However, the OSL-film’s measured dose differences of the IMRT patterns were lower than those of Gafchromic EBT measurements ([-1.6%, 2.1%] versus [-2.9%, 3.6%]). The current OSL-film could be used as a reusable high resolution dosimeter with read-out immediately after

  19. Correcting scan-to-scan response variability for a radiochromic film-based reference dosimetry system

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lewis, David, E-mail: rcfilmconsulting@gmail.com; Devic, Slobodan

    Purpose: In radiochromic film dosimetry systems, measurements are usually obtained from film images acquired on a CCD-based flatbed scanner. The authors investigated factors affecting scan-to-scan response variability leading to increased dose measurement uncertainty. Methods: The authors used flatbed document scanners to repetitively scan EBT3 radiochromic films exposed to doses 0–1000 cGy, together with three neutral density filters and three blue optical filters. Scanning was performed under two conditions: scanner lid closed and scanner lid opened/closed between scans. The authors also placed a scanner in a cold room at 9 °C and later in a room at 22 °C and scanned EBT3 filmsmore » to explore temperature effects. Finally, the authors investigated the effect of altering the distance between the film and the scanner’s light source. Results: Using a measurement protocol to isolate the contribution of the CCD and electronic circuitry of the scanners, the authors found that the standard deviation of response measurements for the EBT3 film model was about 0.17% for one scanner and 0.09% for the second. When the lid of the first scanner was opened and closed between scans, the average scan-to-scan difference of responses increased from 0.12% to 0.27%. Increasing the sample temperature during scanning changed the RGB response values by about −0.17, −0.14, and −0.05%/°C, respectively. Reducing the film-to-light source distance increased the RBG response values about 1.1, 1.3, and 1.4%/mm, respectively. The authors observed that films and film samples were often not flat with some areas up to 8 mm away from the scanner’s glass window. Conclusions: In the absence of measures to deal with the response irregularities, each factor the authors investigated could lead to dose uncertainty >2%. Those factors related to the film-to-light source distance could be particularly impactful since the authors observed many instances where the curl of film samples

  20. Visible absorption properties of radiation exposed XR type-T radiochromic film.

    PubMed

    Butson, Martin J; Cheung, Tsang; Yu, Peter K N

    2004-10-07

    The visible absorption spectra of Gafchromic XR type-T radiochromic film have been investigated to analyse the dosimetry characteristics of the film with visible light densitometers. Common densitometers can use photospectrometry, fluorescent light (broad-band visible), helium neon (632 nm), light emitting diode (LED) or other specific bandwidth spectra. The visible absorption spectra of this film when exposed to photon radiation show peaks at 676 nm and 618 nm at 2 Gy absorbed doses which shift to slightly lower wavelengths (662 nm and 612 nm at 8 Gy absorbed dose) at higher doses. This is similar to previous models of Gafchromic film such as MD-55-2 and HS but XR type-T also includes a large absorption at lower visible wavelengths due to 'yellow' dyes placed within the film to aid with visible recognition of the film exposure level. The yellow dye band pass is produced at approximately 520 nm to 550 nm and absorbs wavelengths lower than this value within the visible spectrum. This accounts for the colour change from yellow to brown through the added absorption in the red wavelengths with radiation exposure. The film produces a relatively high dose sensitivity with up to 0.25 OD units per Gy change at 672 nm at 100 kVp x-ray energy. Variations in dose sensitivity can be achieved by varying wavelength analysis.

  1. Evaluation of exhaled breath temperature (EBT) as a marker and predictor of asthma exacerbation in children and adolescents.

    PubMed

    Wojsyk-Banaszak, Irena; Mikoś, Marcin; Szczepankiewicz, Aleksandra; Wielebska, Alicja; Sobkowiak, Paulina; Kamińska, Aleksandra; Bręborowicz, Anna

    2017-09-01

    Noninvasive and easy-to-use tools to monitor airway inflammation in asthma are needed to maintain disease control, particularly in pediatric population. The aim of the study was to evaluate exhaled breath temperature (EBT) in pediatric respiratory clinic setting. We evaluated 37 children and adolescents with asthma (5-17 years; median: 11 years). The patients were followed up in stable condition and during exacerbations (paired observations in n = 19 subjects). We evaluated medication use, EBT, fractional exhaled nitric oxide (FeNO), spirometry and atopic status of patients. EBT was significantly higher in children with asthma exacerbation {entire group: median [interquartile range (IQR)]: 32.3 [1.1]°C vs. 33.8 [1.7]°C; p < 0.001 and mean ± SD: 33.1 ± 1.0°C vs. 33.6 ± 1.1°C; p = 0.038 for paired observations}. Significant correlation was observed between EBT and FeNO in the entire group (r = 0.22; p = 0.03). No difference was observed in EBT median values in atopic and non-atopic subjects in the entire group (median [IQR]: 32.6 [1.6] vs. 32.7 [2.0]; p = 0.88) and in subgroups. There was no difference in EBT values in patients receiving systemic or inhaled glucocorticosteroids (p = 0.45 and 0.83). There was no significant correlation between EBT and body or room temperature. The only significant predictor of exacerbation in logistic regression model was EBT {aOR = 2.4; 95% [confidence interval (CI)]: 1.4-4.1}. ROC analysis demonstrated applicability of EBT as a marker of asthma exacerbation in children (AUC = 0.748; p < 0.001; cut-off = 33.3°C; sensitivity: 64.3%; specificity: 82.1%). We suggest that EBT may serve as marker and predictor of asthma exacerbation in children. EBT follow-up may be useful in asthma monitoring in children and adolescents.

  2. Multichannel film dosimetry with nonuniformity correction.

    PubMed

    Micke, Andre; Lewis, David F; Yu, Xiang

    2011-05-01

    all the color channels the multichannel method provides the ability to examine the agreement between the color channels. Furthermore, when using calibration data to convert RGB film images to dose using the new method, poor correspondence between the dose calculations for the three color channels provides an important indication that the this new technique enables easy indication in case the dose and calibration films are curve mismatched. The method permit compensation for thickness nonuniformities in the film, increases the signal to noise level, mitigates the lateral dose-dependency of flatbed scanners effect of the calculated dose map and extends the evaluable dose range to 10 cGy-100 Gy. Multichannel dosimetry with radiochromic film like Gafchromic EBT2 is shown to have significant advantages over single channel dosimetry. It is recommended that the dosimetry protocols described be implemented when using this radiochromic film to ensure the best data integrity and dosimetric accuracy.

  3. Could EBT Machines Increase Fruit and Vegetable Purchases at New York City Green Carts?

    PubMed

    Breck, Andrew; Kiszko, Kamila; Martinez, Olivia; Abrams, Courtney; Elbel, Brian

    2017-09-21

    Residents of some low-income neighborhoods have limited access to fresh fruits and vegetables. In 2008, New York City issued new mobile fruit and vegetable cart licenses for neighborhoods with inadequate availability of fresh produce. Some of these carts were equipped with electronic benefit transfer (EBT) machines, allowing them to accept Supplemental Nutrition Assistance Program (SNAP) benefits. This article examines the association between type and quantities of fruits and vegetables purchased from mobile fruit and vegetable vendors and consumer characteristics, including payment method. Customers at 4 produce carts in the Bronx, New York, were surveyed during 3 periods in 2013 and 2014. Survey data, including purchased fruit and vegetable quantities, were analyzed using multivariable negative binomial regressions, with payment method (cash only vs EBT or EBT and cash) as the primary independent variable. Covariates included availability of EBT, vendor, and customer sociodemographic characteristics. A total of 779 adults participated in this study. Shoppers who used SNAP benefits purchased an average of 5.4 more cup equivalents of fruits and vegetables than did shoppers who paid with cash. Approximately 80% of this difference was due to higher quantities of purchased fruits. Expanding access to EBT machines at mobile produce carts may increase purchases of fruits and vegetables from these vendors.

  4. Could EBT Machines Increase Fruit and Vegetable Purchases at New York City Green Carts?

    PubMed Central

    Breck, Andrew; Kiszko, Kamila; Martinez, Olivia; Abrams, Courtney

    2017-01-01

    Introduction Residents of some low-income neighborhoods have limited access to fresh fruits and vegetables. In 2008, New York City issued new mobile fruit and vegetable cart licenses for neighborhoods with inadequate availability of fresh produce. Some of these carts were equipped with electronic benefit transfer (EBT) machines, allowing them to accept Supplemental Nutrition Assistance Program (SNAP) benefits. This article examines the association between type and quantities of fruits and vegetables purchased from mobile fruit and vegetable vendors and consumer characteristics, including payment method. Methods Customers at 4 produce carts in the Bronx, New York, were surveyed during 3 periods in 2013 and 2014. Survey data, including purchased fruit and vegetable quantities, were analyzed using multivariable negative binomial regressions, with payment method (cash only vs EBT or EBT and cash) as the primary independent variable. Covariates included availability of EBT, vendor, and customer sociodemographic characteristics. Results A total of 779 adults participated in this study. Shoppers who used SNAP benefits purchased an average of 5.4 more cup equivalents of fruits and vegetables than did shoppers who paid with cash. Approximately 80% of this difference was due to higher quantities of purchased fruits. Conclusion Expanding access to EBT machines at mobile produce carts may increase purchases of fruits and vegetables from these vendors. PMID:28934080

  5. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chiu-Tsao, S; Chan, M

    Purpose: To study the effect of film scanning orientation for new EBT-XD film using Vidar and Epson scanners, compared with EBT3 films. Methods: The EBT-XD (lot#01081501) and EBT3 films (lot#11031501) were cut into 4cm×7cm pieces and each film piece was uniformly exposed to a 6MV photon beam (15×15cm{sup 2}) at 5 cm depth and 100cm SAD in polystyrene phantom, one at a time. The doses delivered were 0,2.5,5,10,20,30,40 and 50 Gy for EBT-XD films, and 0,0.5,1,2,4,8,12 and 16 Gy for EBT3 films, respectively. All films were scanned at least 16 hours after irradiation, one at a time, in portrait ormore » landscape orientation, using a Vidar DosimetryPro Advantage Red and an Epson 10000XL scanner (red channel). Each film piece was positioned at the center of a frame (36cm×36cm) with a cutout slightly larger than the film size for scanning. The scanned data (71 dpi), were analyzed using ImageJ and the average in the ROI (3cm×3cm) of each film at the center was obtained. Results: Dose response curves (netOD vs. dose) were established and compared with EBT-XD and EBT3 film in portrait or landscape orientation using these two scanners. The portrait to landscape (P/L) ratio for each film/scanner combination was determined. The EBT-XD film is about three times less sensitive than EBT3 film. The netOD is higher for portrait than landscape orientation for all the film/scanner combinations. For EBT-XD and EBT3, the average P/L ratios are 1.02 and 1.03 using Vidar scanner, and 1.03 and 1.05 using Epson scanner (red channel), respectively. Conclusion: The effect of film scanning orientation for new EBT-XD film using Vidar Advantage Red and Epson 10000XL scanners has been studied, in comparison with EBT3 film. The portrait/landscape ratios are the lowest for EBT-XD film using Vidar Advantage Red scanner among the four film/scanner combinations.« less

  6. mEBT: multiple-matching Evidence-based Translator of Murine Genomic Responses for Human Immunity Studies.

    PubMed

    Tae, Donghyun; Seok, Junhee

    2018-05-29

    In this paper, we introduce multiple-matching Evidence-based Translator (mEBT) to discover genomic responses from murine expression data for human immune studies, which are significant in the given condition of mice and likely have similar responses in the corresponding condition of human. mEBT is evaluated over multiple data sets and shows improved inter-species agreement. mEBT is expected to be useful for research groups who use murine models to study human immunity. http://cdal.korea.ac.kr/mebt/. jseok14@korea.ac.kr. Supplementary data are available at Bioinformatics online.

  7. Monaco and film dosimetry of 3D CRT, IMRT and VMAT cases in a realistic pelvic prosthetic phantom

    NASA Astrophysics Data System (ADS)

    Ade, Nicholas; du Plessis, F. C. P.

    2018-04-01

    The dosimetry of patients with metallic hip implants during irradiation of pelvic lesions is challenging due to dose distortions caused by implants. This work presents a dosimetric comparison of various multi-field photon-beam dose distributions in the presence of unilateral hip titanium prosthesis (UHTiP) embedded in a unique pelvic phantom made out of water-equivalent nylon slices. The impact of the UHTiP on the accuracy of dose calculations from a Monaco TPS (treatment planning system) using the X-ray voxel Monte Carlo (XVMC) algorithm was benchmarked against measured dose data using Gafchromic EBT3 film. Multi-field beam arrangements including a 4-field box, 5-field 3DCRT (three-dimensional conformal radiation therapy), 6-field IMRT (intensity modulated radiation therapy) and a single-arc VMAT (volumetric modulated arc therapy) plan were set up for 6 MV and 15 MV beams. These plans were generated for the pelvic phantom that contains the prosthesis with film inserted. Compared to Monaco TPS dose calculations, film measurements showed enhanced dose in the prosthesis which was not predicted by Monaco due to its limitation in relative density assignment. The enhanced prosthesis dose increased with increase in beam energy and decreased with the complexity of the treatment plans, with VMAT giving the least escalated dose. The dose increased between 5% and 19% for 6 MV and between 6% and 21% for 15 MV. A gamma index analysis showed that 70-92% of dose points (excluding the prosthesis) were within 3% discrepancy. Increasing the number of treatment fields increases target dose coverage and improves the agreement between film and Monaco. When the relative electron density (RED) in the prosthesis was varied between 3.72 and 15 the dose discrepancy between film and Monaco increased from 30% to 57% for 6 MV and from 30% to 50% for 15 MV. The study indicates that beam weights for fields that pass through the prosthesis should be minimised and its RED must be correct for

  8. SU-C-BRC-05: Monte Carlo Calculations to Establish a Simple Relation of Backscatter Dose Enhancement Around High-Z Dental Alloy to Its Atomic Number

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Utsunomiya, S; Kushima, N; Katsura, K

    Purpose: To establish a simple relation of backscatter dose enhancement around a high-Z dental alloy in head and neck radiation therapy to its average atomic number based on Monte Carlo calculations. Methods: The PHITS Monte Carlo code was used to calculate dose enhancement, which is quantified by the backscatter dose factor (BSDF). The accuracy of the beam modeling with PHITS was verified by comparing with basic measured data namely PDDs and dose profiles. In the simulation, a high-Z alloy of 1 cm cube was embedded into a tough water phantom irradiated by a 6-MV (nominal) X-ray beam of 10 cmmore » × 10 cm field size of Novalis TX (Brainlab). The ten different materials of high-Z alloys (Al, Ti, Cu, Ag, Au-Pd-Ag, I, Ba, W, Au, Pb) were considered. The accuracy of calculated BSDF was verified by comparing with measured data by Gafchromic EBT3 films placed at from 0 to 10 mm away from a high-Z alloy (Au-Pd-Ag). We derived an approximate equation to determine the relation of BSDF and range of backscatter to average atomic number of high-Z alloy. Results: The calculated BSDF showed excellent agreement with measured one by Gafchromic EBT3 films at from 0 to 10 mm away from the high-Z alloy. We found the simple linear relation of BSDF and range of backscatter to average atomic number of dental alloys. The latter relation was proven by the fact that energy spectrum of backscatter electrons strongly depend on average atomic number. Conclusion: We found a simple relation of backscatter dose enhancement around high-Z alloys to its average atomic number based on Monte Carlo calculations. This work provides a simple and useful method to estimate backscatter dose enhancement from dental alloys and corresponding optimal thickness of dental spacer to prevent mucositis effectively.« less

  9. Determination of MLC model parameters for Monaco using commercial diode arrays.

    PubMed

    Kinsella, Paul; Shields, Laura; McCavana, Patrick; McClean, Brendan; Langan, Brian

    2016-07-08

    Multileaf collimators (MLCs) need to be characterized accurately in treatment planning systems to facilitate accurate intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT). The aim of this study was to examine the use of MapCHECK 2 and ArcCHECK diode arrays for optimizing MLC parameters in Monaco X-ray voxel Monte Carlo (XVMC) dose calculation algorithm. A series of radiation test beams designed to evaluate MLC model parameters were delivered to MapCHECK 2, ArcCHECK, and EBT3 Gafchromic film for comparison. Initial comparison of the calculated and ArcCHECK-measured dose distributions revealed it was unclear how to change the MLC parameters to gain agreement. This ambiguity arose due to an insufficient sampling of the test field dose distributions and unexpected discrepancies in the open parts of some test fields. Consequently, the XVMC MLC parameters were optimized based on MapCHECK 2 measurements. Gafchromic EBT3 film was used to verify the accuracy of MapCHECK 2 measured dose distributions. It was found that adjustment of the MLC parameters from their default values resulted in improved global gamma analysis pass rates for MapCHECK 2 measurements versus calculated dose. The lowest pass rate of any MLC-modulated test beam improved from 68.5% to 93.5% with 3% and 2 mm gamma criteria. Given the close agreement of the optimized model to both MapCHECK 2 and film, the optimized model was used as a benchmark to highlight the relatively large discrepancies in some of the test field dose distributions found with ArcCHECK. Comparison between the optimized model-calculated dose and ArcCHECK-measured dose resulted in global gamma pass rates which ranged from 70.0%-97.9% for gamma criteria of 3% and 2 mm. The simple square fields yielded high pass rates. The lower gamma pass rates were attributed to the ArcCHECK overestimating the dose in-field for the rectangular test fields whose long axis was parallel to the long axis of the Arc

  10. Study on the coloration response of a radiochromic film to MeV cluster ion beams

    NASA Astrophysics Data System (ADS)

    Yuri, Yosuke; Narumi, Kazumasa; Chiba, Atsuya; Hirano, Yoshimi; Saitoh, Yuichi

    2017-11-01

    A radiochromic film, Gafchromic HD-V2, is applied to a possible method of measuring a two-dimensional (2D) spatial profile of MeV cluster ion beams. The coloration responses of the HD-V2 film to MeV carbon and gold cluster ion beams are experimentally investigated since some cluster effect may appear. The degree of the film coloration is quantified as a change in optical density (OD) by reading the films with an image scanner for high-resolution measurement of the 2D beam profile. The OD response of HD-V2 is characterized as a function of the ion and atom fluence for comparison. The dependences of the OD response on the cluster size, kinetic energy, and ion species are discussed. It is found that the sensitivity of the OD change is reduced when the cluster size is large. The beam profile of MeV cluster ion beams delivered from the tandem accelerator in TIARA is characterized from the measurement result using HD-V2 films. The present results show that the use of the Gafchromic HD-V2 film is suitable for the detail beam profile measurement of MeV cluster ions, especially C60 ions, whose available intensity is rather low in comparison with that of monatomic ion beams.

  11. SU-E-T-205: Improving Quality Assurance of HDR Brachytherapy: Verifying Agreement Between Planned and Delivered Dose Distributions Using DICOM RTDose and Advanced Film Dosimetry

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Palmer, A L; University of Surrey, Guildford, Surrey; Bradley, D A

    Purpose: HDR brachytherapy is undergoing significant development, and quality assurance (QA) checks must keep pace. Current recommendations do not adequately verify delivered against planned dose distributions: This is particularly relevant for new treatment planning system (TPS) calculation algorithms (non TG-43 based), and an era of significant patient-specific plan optimisation. Full system checks are desirable in modern QA recommendations, complementary to device-centric individual tests. We present a QA system incorporating TPS calculation, dose distribution export, HDR unit performance, and dose distribution measurement. Such an approach, more common in external beam radiotherapy, has not previously been reported in the literature for brachytherapy.more » Methods: Our QA method was tested at 24 UK brachytherapy centres. As a novel approach, we used the TPS DICOM RTDose file export to compare planned dose distribution with that measured using Gafchromic EBT3 films placed around clinical brachytherapy treatment applicators. Gamma analysis was used to compare the dose distributions. Dose difference and distance to agreement were determined at prescription Point A. Accurate film dosimetry was achieved using a glass compression plate at scanning to ensure physically-flat films, simultaneous scanning of known dose films with measurement films, and triple-channel dosimetric analysis. Results: The mean gamma pass rate of RTDose compared to film-measured dose distributions was 98.1% at 3%(local), 2 mm criteria. The mean dose difference, measured to planned, at Point A was -0.5% for plastic treatment applicators and -2.4% for metal applicators, due to shielding not accounted for in TPS. The mean distance to agreement was 0.6 mm. Conclusion: It is recommended to develop brachytherapy QA to include full-system verification of agreement between planned and delivered dose distributions. This is a novel approach for HDR brachytherapy QA. A methodology using advanced film

  12. Proof of principle study of the use of a CMOS active pixel sensor for proton radiography.

    PubMed

    Seco, Joao; Depauw, Nicolas

    2011-02-01

    Proof of principle study of the use of a CMOS active pixel sensor (APS) in producing proton radiographic images using the proton beam at the Massachusetts General Hospital (MGH). A CMOS APS, previously tested for use in s-ray radiation therapy applications, was used for proton beam radiographic imaging at the MGH. Two different setups were used as a proof of principle that CMOS can be used as proton imaging device: (i) a pen with two metal screws to assess spatial resolution of the CMOS and (ii) a phantom with lung tissue, bone tissue, and water to assess tissue contrast of the CMOS. The sensor was then traversed by a double scattered monoenergetic proton beam at 117 MeV, and the energy deposition inside the detector was recorded to assess its energy response. Conventional x-ray images with similar setup at voltages of 70 kVp and proton images using commercial Gafchromic EBT 2 and Kodak X-Omat V films were also taken for comparison purposes. Images were successfully acquired and compared to x-ray kVp and proton EBT2/X-Omat film images. The spatial resolution of the CMOS detector image is subjectively comparable to the EBT2 and Kodak X-Omat V film images obtained at the same object-detector distance. X-rays have apparent higher spatial resolution than the CMOS. However, further studies with different commercial films using proton beam irradiation demonstrate that the distance of the detector to the object is important to the amount of proton scatter contributing to the proton image. Proton images obtained with films at different distances from the source indicate that proton scatter significantly affects the CMOS image quality. Proton radiographic images were successfully acquired at MGH using a CMOS active pixel sensor detector. The CMOS demonstrated spatial resolution subjectively comparable to films at the same object-detector distance. Further work will be done in order to establish the spatial and energy resolution of the CMOS detector for protons. The

  13. Proof of principle study of the use of a CMOS active pixel sensor for proton radiography

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Seco, Joao; Depauw, Nicolas

    2011-02-15

    Purpose: Proof of principle study of the use of a CMOS active pixel sensor (APS) in producing proton radiographic images using the proton beam at the Massachusetts General Hospital (MGH). Methods: A CMOS APS, previously tested for use in s-ray radiation therapy applications, was used for proton beam radiographic imaging at the MGH. Two different setups were used as a proof of principle that CMOS can be used as proton imaging device: (i) a pen with two metal screws to assess spatial resolution of the CMOS and (ii) a phantom with lung tissue, bone tissue, and water to assess tissuemore » contrast of the CMOS. The sensor was then traversed by a double scattered monoenergetic proton beam at 117 MeV, and the energy deposition inside the detector was recorded to assess its energy response. Conventional x-ray images with similar setup at voltages of 70 kVp and proton images using commercial Gafchromic EBT 2 and Kodak X-Omat V films were also taken for comparison purposes. Results: Images were successfully acquired and compared to x-ray kVp and proton EBT2/X-Omat film images. The spatial resolution of the CMOS detector image is subjectively comparable to the EBT2 and Kodak X-Omat V film images obtained at the same object-detector distance. X-rays have apparent higher spatial resolution than the CMOS. However, further studies with different commercial films using proton beam irradiation demonstrate that the distance of the detector to the object is important to the amount of proton scatter contributing to the proton image. Proton images obtained with films at different distances from the source indicate that proton scatter significantly affects the CMOS image quality. Conclusion: Proton radiographic images were successfully acquired at MGH using a CMOS active pixel sensor detector. The CMOS demonstrated spatial resolution subjectively comparable to films at the same object-detector distance. Further work will be done in order to establish the spatial and energy resolution

  14. Objective coding of content and techniques in workplace-based supervision of an EBT in public mental health.

    PubMed

    Dorsey, Shannon; Kerns, Suzanne E U; Lucid, Leah; Pullmann, Michael D; Harrison, Julie P; Berliner, Lucy; Thompson, Kelly; Deblinger, Esther

    2018-01-24

    Workplace-based clinical supervision as an implementation strategy to support evidence-based treatment (EBT) in public mental health has received limited research attention. A commonly provided infrastructure support, it may offer a relatively cost-neutral implementation strategy for organizations. However, research has not objectively examined workplace-based supervision of EBT and specifically how it might differ from EBT supervision provided in efficacy and effectiveness trials. Data come from a descriptive study of supervision in the context of a state-funded EBT implementation effort. Verbal interactions from audio recordings of 438 supervision sessions between 28 supervisors and 70 clinicians from 17 public mental health organizations (in 23 offices) were objectively coded for presence and intensity coverage of 29 supervision strategies (16 content and 13 technique items), duration, and temporal focus. Random effects mixed models estimated proportion of variance in content and techniques attributable to the supervisor and clinician levels. Interrater reliability among coders was excellent. EBT cases averaged 12.4 min of supervision per session. Intensity of coverage for EBT content varied, with some discussed frequently at medium or high intensity (exposure) and others infrequently discussed or discussed only at low intensity (behavior management; assigning/reviewing client homework). Other than fidelity assessment, supervision techniques common in treatment trials (e.g., reviewing actual practice, behavioral rehearsal) were used rarely or primarily at low intensity. In general, EBT content clustered more at the clinician level; different techniques clustered at either the clinician or supervisor level. Workplace-based clinical supervision may be a feasible implementation strategy for supporting EBT implementation, yet it differs from supervision in treatment trials. Time allotted per case is limited, compressing time for EBT coverage. Techniques that

  15. SU-E-T-149: Brachytherapy Patient Specific Quality Assurance for a HDR Vaginal Cylinder Case

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Barbiere, J; Napoli, J; Ndlovu, A

    2015-06-15

    Purpose: Commonly Ir-192 HDR treatment planning system commissioning is only based on a single absolute measurement of source activity supplemented by tabulated parameters for multiple factors without independent verification that the planned distribution corresponds to the actual delivered dose. The purpose on this work is to present a methodology using Gafchromic film with a statistically valid calibration curve that can be used to validate clinical HDR vaginal cylinder cases by comparing the calculated plan dose distribution in a plane with the corresponding measured planar dose. Methods: A vaginal cylinder plan was created with Oncentra treatment planning system. The 3D dosemore » matrix was exported to a Varian Eclipse work station for convenient extraction of a 2D coronal dose plane corresponding to the film position. The plan was delivered with a sheet of Gafchromic EBT3 film positioned 1mm from the catheter using an Ir-192 Nucletron HDR source. The film was then digitized with an Epson 10000 XL color scanner. Film analysis is performed with MatLab imaging toolbox. A density to dose calibration curve was created using TG43 formalism for a single dwell position exposure at over 100 points for statistical accuracy. The plan and measured film dose planes were registered using a known dwell position relative to four film marks. The plan delivered 500 cGy to points 2 cm from the sources. Results: The distance to agreement of the 500 cGy isodose between the plan and film measurement laterally was 0.5 mm but can be as much as 1.5 mm superior and inferior. The difference between the computed plan dose and film measurement was calculated per pixel. The greatest errors up to 50 cGy are near the apex. Conclusion: The methodology presented will be useful to implement more comprehensive quality assurance to verify patient-specific dose distributions.« less

  16. Increasing clinicians' EBT exploration and preparation behavior in youth mental health services by changing organizational culture with ARC.

    PubMed

    Glisson, Charles; Williams, Nathaniel J; Hemmelgarn, Anthony; Proctor, Enola; Green, Philip

    2016-01-01

    Clinician EBT exploration and preparation behavior is essential to the ongoing implementation of new EBTs. This study examined the effect of the ARC organizational intervention on clinician EBT exploration and preparation behavior and assessed the mediating role of organizational culture as a linking mechanism. Fourteen community mental health agencies that serve youth in a major Midwestern metropolis along with 475 clinicians who worked in those agencies, were randomly assigned to either the three-year ARC intervention or control condition. Organizational culture was assessed with the Organizational Social Context (OSC) measure at baseline and follow-up. EBT exploration and preparation behavior was measured as clinician participation in nine separate community EBT workshops held over a three-year period. There was 69 percent greater odds (OR = 1.69, p < .003) of clinicians in the ARC condition (versus control) attending each subsequent workshop. Workshop attendance in the control group remained under two percent and declined over three years while attendance in the ARC condition grew from 3.6 percent in the first workshop to 12 percent in the ninth and final workshop. Improvement in proficient organizational culture mediated the positive effect of the ARC intervention on clinicians' workshop attendance (a × b = .21; 95% CI:LL = .05, UL = .40). Organizational interventions that create proficient mental health agency cultures can increase clinician EBT exploration and preparation behavior that is essential to the ongoing implementation of new EBTs in community youth mental health settings. Copyright © 2015 Elsevier Ltd. All rights reserved.

  17. SU-G-201-06: Directional Low-Dose Rate Brachytherapy: Determination of the TG-43 Dose-Rate Constant Analog for a New Pd-103 Source

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Aima, M; Culberson, W; Hammer, C

    Purpose: The aim of this work is to determine the TG-43 dose-rate constant analog for a new directional low-dose rate brachytherapy source based on experimental methods and comparison to Monte Carlo simulations. The CivaSheet™ is a new commercially available planar source array comprised of a variable number of discrete directional source elements called “CivaDots”. Given the directional nature and non-conventional design of the source, modifications to the AAPM TG-43 protocol for dosimetry are required. As a result, various parameters of the TG-43 dosimetric formalism have to be adapted to accommodate this source. This work focuses on the dose-rate constant analogmore » determination for a CivaDot. Methods: Dose to water measurements of the CivaDot were performed in a polymethyl methacrylate phantom (20×20×12 cm{sup 3}) using thermoluminescent dosimeters (TLDs) and Gafchromic EBT3 film. The source was placed in the center of the phantom, and nine TLD micro-cubes were irradiated along its central axis at a distance of 1 cm. For the film measurements, the TLDs were substituted by a (3×3) cm{sup 2} EBT3 film. Primary air-kerma strength measurements of the source were performed using a variable-aperture free-air chamber. Finally, the source was modeled using the Monte Carlo N-Particle Transport Code 6. Results: Dose-rate constant analog observed for a total of eight CivaDots using TLDs and five CivaDots using EBT3 film was within ±7.0% and ±2.9% of the Monte Carlo predicted value respectively. The average difference observed was −4.8% and −0.1% with a standard deviation of 1.7% and 2.1% for the TLD and the film measurements respectively, which are both within the comparison uncertainty. Conclusion: A preliminary investigation to determine the doserate constant analog for a CivaDot was conducted successfully with good agreement between experimental and Monte Carlo based methods. This work will aid in the eventual realization of a clinically

  18. On the Use of Optically Stimulated Luminescent Dosimeter for Surface Dose Measurement during Radiotherapy

    PubMed Central

    Yusof, Fasihah Hanum; Ung, Ngie Min; Wong, Jeannie Hsiu Ding; Jong, Wei Loong; Ath, Vannyat; Phua, Vincent Chee Ee; Heng, Siew Ping; Ng, Kwan Hoong

    2015-01-01

    This study was carried out to investigate the suitability of using the optically stimulated luminescence dosimeter (OSLD) in measuring surface dose during radiotherapy. The water equivalent depth (WED) of the OSLD was first determined by comparing the surface dose measured using the OSLD with the percentage depth dose at the buildup region measured using a Markus ionization chamber. Surface doses were measured on a solid water phantom using the OSLD and compared against the Markus ionization chamber and Gafchromic EBT3 film measurements. The effect of incident beam angles on surface dose was also studied. The OSLD was subsequently used to measure surface dose during tangential breast radiotherapy treatments in a phantom study and in the clinical measurement of 10 patients. Surface dose to the treated breast or chest wall, and on the contralateral breast were measured. The WED of the OSLD was found to be at 0.4 mm. For surface dose measurement on a solid water phantom, the Markus ionization chamber measured 15.95% for 6 MV photon beam and 12.64% for 10 MV photon beam followed by EBT3 film (23.79% and 17.14%) and OSLD (37.77% and 25.38%). Surface dose increased with the increase of the incident beam angle. For phantom and patient breast surface dose measurement, the response of the OSLD was higher than EBT3 film. The in-vivo measurements were also compared with the treatment planning system predicted dose. The OSLD measured higher dose values compared to dose at the surface (Hp(0.0)) by a factor of 2.37 for 6 MV and 2.01 for 10 MV photon beams, respectively. The measurement of absorbed dose at the skin depth of 0.4 mm by the OSLD can still be a useful tool to assess radiation effects on the skin dermis layer. This knowledge can be used to prevent and manage potential acute skin reaction and late skin toxicity from radiotherapy treatments. PMID:26052690

  19. An investigation into factors affecting the precision of CT radiation dose profile width measurements using radiochromic films

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Li, Baojun, E-mail: Baojunli@bu.edu; Behrman, Richard H.

    Purpose: To investigate the impact of x-ray beam energy, exposure intensity, and flat-bed scanner uniformity and spatial resolution on the precision of computed tomography (CT) beam width measurements using Gafchromic XR-QA2 film and an off-the-shelf document scanner. Methods: Small strips of Gafchromic film were placed at isocenter in a CT scanner and exposed at various x-ray beam energies (80–140 kVp), exposure levels (50–400 mA s), and nominal beam widths (1.25, 5, and 10 mm). The films were scanned in reflection mode on a Ricoh MP3501 flat-bed document scanner using several spatial resolution settings (100 to 400 dpi) and at differentmore » locations on the scanner bed. Reflection measurements were captured in digital image files and radiation dose profiles generated by converting the image pixel values to air kerma through film calibration. Beam widths were characterized by full width at half maximum (FWHM) and full width at tenth maximum (FWTM) of dose profiles. Dependences of these parameters on the above factors were quantified in percentage change from the baselines. Results: The uncertainties in both FWHM and FWTM caused by varying beam energy, exposure level, and scanner uniformity were all within 4.5% and 7.6%, respectively. Increasing scanner spatial resolution significantly increased the uncertainty in both FWHM and FWTM, with FWTM affected by almost 8 times more than FWHM (48.7% vs 6.5%). When uncalibrated dose profiles were used, FWHM and FWTM were over-estimated by 11.6% and 7.6%, respectively. Narrower beam width appeared more sensitive to the film calibration than the wider ones (R{sup 2} = 0.68 and 0.85 for FWHM and FWTM, respectively). The global and maximum local background variations of the document scanner were 1.2%. The intrinsic film nonuniformity for an unexposed film was 0.3%. Conclusions: Measurement of CT beam widths using Gafchromic XR-QA2 films is robust against x-ray energy, exposure level, and scanner uniformity. With proper

  20. A new correction method serving to eliminate the parabola effect of flatbed scanners used in radiochromic film dosimetry.

    PubMed

    Poppinga, D; Schoenfeld, A A; Doerner, K J; Blanck, O; Harder, D; Poppe, B

    2014-02-01

    The purpose of this study is the correction of the lateral scanner artifact, i.e., the effect that, on a large homogeneously exposed EBT3 film, a flatbed scanner measures different optical densities at different positions along the x axis, the axis parallel to the elongated light source. At constant dose, the measured optical density profiles along this axis have a parabolic shape with significant dose dependent curvature. Therefore, the effect is shortly called the parabola effect. The objective of the algorithm developed in this study is to correct for the parabola effect. Any optical density measured at given position x is transformed into the equivalent optical density c at the apex of the parabola and then converted into the corresponding dose via the calibration of c versus dose. For the present study EBT3 films and an Epson 10000XL scanner including transparency unit were used for the analysis of the parabola effect. The films were irradiated with 6 MV photons from an Elekta Synergy accelerator in a RW3 slab phantom. In order to quantify the effect, ten film pieces with doses graded from 0 to 20.9 Gy were sequentially scanned at eight positions along the x axis and at six positions along the z axis (the movement direction of the light source) both for the portrait and landscape film orientations. In order to test the effectiveness of the new correction algorithm, the dose profiles of an open square field and an IMRT plan were measured by EBT3 films and compared with ionization chamber and ionization chamber array measurement. The parabola effect has been numerically studied over the whole measuring field of the Epson 10000XL scanner for doses up to 20.9 Gy and for both film orientations. The presented algorithm transforms any optical density at position x into the equivalent optical density that would be measured at the same dose at the apex of the parabola. This correction method has been validated up to doses of 5.2 Gy all over the scanner bed with 2D dose

  1. Can a commercial gel dosimetry system be used to verify stereotactic spinal radiotherapy treatment dose distributions?

    NASA Astrophysics Data System (ADS)

    Kairn, T.; Asena, A.; Crowe, S. B.; Livingstone, A.; Papworth, D.; Smith, S.; Sutherland, B.; Sylvander, S.; Franich, R. D.; Trapp, J. V.

    2017-05-01

    This study investigated the use of the TruView xylenol-orange-based gel and VISTA optical CT scanner (both by Modus Medical Inc, London, Canada), for use in verifying the accuracy of planned dose distributions for hypo-fractionated (stereotactic) vertebral treatments. Gel measurements were carried out using three stereotactic vertebral treatments and compared with planned doses calculated using the Eclipse treatment planning system (Varian Medical Systems, Palo Alto, USA) as well as with film measurements made using Gafchromic EBT3 film (Ashland Inc, Covington, USA), to investigate the accuracy of the gel system. The gel was calibrated with reference to a moderate-dose gradient region in one of the gel samples. Generally, the gel measurements were able to approximate the close agreement between the doses calculated by the treatment planning system and the doses measured using film (which agreed with each other within 2%), despite lower resolution and bit depth. Poorer agreement was observed when the dose delivered to the gel exceeded the range of doses delivered in the calibration region. This commercial gel dosimetry system may be used to verify hypo-fractionated treatments of vertebral targets, although separate gel calibration measurements are recommended.

  2. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Saur, Sigrun; Frengen, Jomar; Department of Oncology and Radiotherapy, St. Olavs University Hospital, N-7006 Trondheim

    Film dosimetry using radiochromic EBT film in combination with a flatbed charge coupled device scanner is a useful method both for two-dimensional verification of intensity-modulated radiation treatment plans and for general quality assurance of treatment planning systems and linear accelerators. Unfortunately, the response over the scanner area is nonuniform, and when not corrected for, this results in a systematic error in the measured dose which is both dose and position dependent. In this study a novel method for background correction is presented. The method is based on the subtraction of a correction matrix, a matrix that is based on scansmore » of films that are irradiated to nine dose levels in the range 0.08-2.93 Gy. Because the response of the film is dependent on the film's orientation with respect to the scanner, correction matrices for both landscape oriented and portrait oriented scans were made. In addition to the background correction method, a full dose uncertainty analysis of the film dosimetry procedure was performed. This analysis takes into account the fit uncertainty of the calibration curve, the variation in response for different film sheets, the nonuniformity after background correction, and the noise in the scanned films. The film analysis was performed for film pieces of size 16x16 cm, all with the same lot number, and all irradiations were done perpendicular onto the films. The results show that the 2-sigma dose uncertainty at 2 Gy is about 5% and 3.5% for landscape and portrait scans, respectively. The uncertainty gradually increases as the dose decreases, but at 1 Gy the 2-sigma dose uncertainty is still as good as 6% and 4% for landscape and portrait scans, respectively. The study shows that film dosimetry using GafChromic EBT film, an Epson Expression 1680 Professional scanner and a dedicated background correction technique gives precise and accurate results. For the purpose of dosimetric verification, the calculated dose

  3. 64 slice MDCT generally underestimates coronary calcium scores as compared to EBT: A phantom study

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Greuter, M. J. W.; Dijkstra, H.; Groen, J. M.

    The objective of our study was the determination of the influence of the sequential and spiral acquisition modes on the concordance and deviation of the calcium score on 64-slice multi-detector computed tomography (MDCT) scanners in comparison to electron beam tomography (EBT) as the gold standard. Our methods and materials were an anthropomorphic cardio CT phantom with different calcium inserts scanned in sequential and spiral acquisition modes on three identical 64-slice MDCT scanners of manufacturer A and on three identical 64-slice MDCT scanners of manufacturer B and on an EBT system. Every scan was repeated 30 times with and 15 timesmore » without a small random variation in the phantom position for both sequential and spiral modes. Significant differences were observed between EBT and 64-slice MDCT data for all inserts, both acquisition modes, and both manufacturers of MDCT systems. High regression coefficients (0.90-0.98) were found between the EBT and 64-slice MDCT data for both scoring methods and both systems with high correlation coefficients (R{sup 2}>0.94). System A showed more significant differences between spiral and sequential mode than system B. Almost no differences were observed in scanners of the same manufacturer for the Agatston score and no differences for the Volume score. The deviations of the Agatston and Volume scores showed regression dependencies approximately equal to the square root of the absolute score. The Agatston and Volume scores obtained with 64-slice MDCT imaging are highly correlated with EBT-obtained scores but are significantly underestimated (-10% to -2%) for both sequential and spiral acquisition modes. System B is more independent of acquisition mode to calcium score than system A. The Volume score shows no intramanufacturer dependency and its use is advocated versus the Agatston score. Using the same cut points for MDCT-based calcium scores as for EBT-based calcium scores can result in classifying individuals into a

  4. Farmers' Market Manager's Level of Communication and Influence on Electronic Benefits Transfer (EBT) Adoption at Midwest Farmers' Markets.

    PubMed

    Hasin, Afroza; Smith, Sylvia

    2018-01-01

    To understand market managers' level of communication and use of technology that might influence decision to adopt Electronic Benefits Transfer (EBT) at farmers' markets. Cross-sectional study using the Theory of Diffusion of Innovation. Electronic survey administered in midwest states of Illinois, Michigan, and Wisconsin. Farmers' market managers in Illinois, Michigan, and Wisconsin. Information on EBT adoption, market managers' communication, and technology use. Binary logistic regression analysis with EBT adoption as the dependent variable and frequency of technology use, partnership with organizations, farmers' market association (FMA) membership, Facebook page and Web site for the market, and primary source of information as independent variables. Chi-square tests and ANOVA were used to compare states and adopter categories. Logistic regression results showed that the odds of adopting EBT was 7.5 times higher for markets that had partnership with other organizations. Compared with non-adopters, a significantly greater number of early adopters had partnership, FMA membership, and a Facebook page and Web site for market, and reported to a board of directors. Markets that had partnership, FMA membership, a Facebook page and Web site, and mandatory reporting to a board of directors were important factors that influenced EBT adoption at midwest farmers' markets. Copyright © 2017 Society for Nutrition Education and Behavior. Published by Elsevier Inc. All rights reserved.

  5. Characterization of a 2.5 MV inline portal imaging beam

    PubMed Central

    Owen, Jennifer; Eduardo Villarreal‐Barajas, J.; Khan, Rao F.H.

    2016-01-01

    A new megavoltage (MV) energy was recently introduced on Varian TrueBeam linear accelerators for imaging applications. This work describes the experimental characterization of a 2.5 MV inline portal imaging beam for commissioning, routine clinical use, and quality assurance purposes. The beam quality of the 2.5 MV beam was determined by measuring a percent depth dose, PDD, in water phantom for 10×10 cm2 field at source‐to‐surface distance 100 cm with a CC13 ion chamber, plane parallel Markus chamber, and GafChromic EBT3 film. Absolute dosimetric output calibration of the beam was performed using a traceable calibrated ionization chamber, following the AAPM Task Group 51 procedure. EBT3 film measurements were also performed to measure entrance dose. The output stability of the imaging beam was monitored for five months. Coincidence of 2.5 MV imaging beam with 6 MV therapy beam was verified with hidden‐target cubic phantom. Image quality was studied using the Leeds and QC3 phantom. The depth of maximum dose, dmax, and percent dose at 10 cm depth were, respectively, 5.7 mm and 51.7% for CC13, 6.1 mm and 51.9% for Markus chamber, and 5.1 mm and 51.9% for EBT3 film. The 2.5 MV beam quality is slightly inferior to that of a  60Co teletherapy beam; however, an estimated kQ of 1.00 was used for output calibration purposes. The beam output was found to be stable to within 1% over a five‐month period. The relative entrance dose as measured with EBT3 films was 63%, compared to 23% for a clinical 6 MV beam for a 10×10 cm2 field. Overall coincidence of the 2.5 MV imaging beam with the 6 MV clinical therapy beam was within 0.2 mm. Image quality results for two commonly used imaging phantoms were superior for the 2.5 MV beam when compared to the conventional 6 MV beam. The results from measurements on two TrueBeam accelerators show that 2.5 MV imaging beam is slightly softer than a therapeutic  60Co beam, it provides superior image quality than a 6 MV therapy

  6. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ferreira, C; Ahmad, S; Firestone, B

    Purpose: To compare dosimetrically three plan calculation systems (Plato, Varian Brachytherapy, and in-house-made Excel) available for I-125 COMS eye plaque treatment with measurement. Methods: All systems assume homogeneous media and calculations are based on a three-dimensional Cartesian coordinates, Plato and Brachytherapy Planning are based on AAPM TG-43 and the in-house Excel program only on inverse square corrections. Doses at specific depths were measured with EBT3 Gafchromic film from a fully loaded and a partially loaded 16 mm plaque (13 and 8 seeds respectively, I-125, model 6711 GE, Oncura). Measurements took place in a water tank, utilizing solid water blocks andmore » a 3D-printed plaque holder. Taking advantage that gafchromic film has low energy dependence, a dose step wedge was delivered with 6 MV photon beam from a Varian 2100 EX linac for calibration. The gray-scale to dose in cGy was obtained with an Epson Expression 10000 XL scanner in the green channel. Treatment plans were generated for doses of 2200 cGy to a depth of 7 mm, and measurements were taken on a sagittal plane. Results: The calculated dose at the prescription point was 2242, 2344, and 2211 cGy with Excel, Brachyvision and Plato respectively for a fully loaded plaque, for the partially loaded plaque the doses were 2266, 2477, and 2193 cGy respectively. At 5 mm depth the doses for Brachyvision and Plato were comparable (3399 and 3267 cGy respectively), however, the measured dose in film was 3180 cGy which was lower by as much as 6.4% in the fully loaded plaque and 7.6% in the partially loaded plaque. Conclusion: Careful methodology and calibration are essential when measuring doses at specific depth due to the sensitivity and rapid dose fall off of I-125.« less

  7. SU-E-CAMPUS-T-05: Preliminary Results On a 2D Dosimetry System Based On the Optically Stimulated Luminescence of Al2O3

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ahmed, M; Eller, S; Yukihara, E

    2014-06-15

    Purpose: To develop a precise 2D dose mapping technique based on the optically stimulated luminescence (OSL) from Al{sub 2}O{sub 3} films for medical applications. Methods: A 2D laser scanning reader was developed using fast F{sup +}-center (lifetime of <7 ns) and slow F-center (lifetime of 35 ms) OSL emission from newly developed Al{sub 2}O{sub 3} films (Landauer Inc.). An algorithm was developed to correct images for both material and system properties. Since greater contribution of the F??-center emission in the recorded signal increases the readout efficiency and robustness of image corrections, Al{sub 2}O{sub 3}:C,Mg film samples are being investigated inmore » addition to Al{sub 2}O{sub 3}:C samples. Preliminary investigations include exposure of the films to a 6 MV photon beam at 10 cm depth in solid water phantom with an SSD of 100 cm, using a 10 cm × 10 cm flat field or a 4 cm × 4 cm field with a 60° wedge filter. Kodak EDR2 radiographic film and EBT2 Gafchromic film were also exposed for comparison. Results: The results indicate that the algorithm is able to correct images and calculate 2D dose. For the wedge field irradiation, the calculated dose at the center of the field was 0.9 Gy for Al{sub 2}O{sub 3}:C and 0.87 Gy for Al{sub 2}O{sub 3}:C,Mg, whereas, the delivered dose was 0.95 Gy. A good qualitative agreement of the dose profiles was obtained between the OSL films and EDR2 and EBT2 films. Laboratory tests using a beta source suggest that a large dynamic range (10{sup −2}−10{sup 2} Gy) can be achieved using this technique. Conclusion: A 2D dosimetry system and an in-house image correction algorithm were developed for 2D film dosimetry in medical applications. The system is in the preliminary stage of development, but the data demonstrates the feasibility of this approach. This work was supported by Landauer, Inc.« less

  8. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lee, S; Zheng, Y; Albani, D

    Purpose: To reduce internal target volume (ITV), respiratory management is a must in imaging and treatment for lung, liver, and breast cancers. We investigated the dosimetric accuracy of VMAT treatment delivery with a Response™ gating system linked to linear accelerator. Methods: The Response™ gating module designed to directly control radiation beam by breath-holding with a ABC system (Elekta AB, Stockholm, Sweden) was tested for VMAT treatments. Seven VMAT plans including three conventional and four stereotactic body radiotherapy (SBRT) cases were evaluated. Each plan was composed of two or four arcs of 6MV radiation beam with prescribed dose ranged from 1.8more » to 9 Gy per fraction. Each plan was delivered continuously without gating and delivered with multiple interruptions by the ResponseTM gating module with a 20 or 30 second breath-holding period. MapCheck2 and Gafchromic EBT3 films sandwiched in MapPHAN were used to measure the delivered dose with and without gating. Films were scanned on a flatbed color scanner, and red channel was extracted for film dosimetry. Gamma analysis was performed to analyze the dosimetrical accuracy of the radiation delivery with gating. Results: The measured doses with gating remarkably agree with the planned dose distributions in the results of gamma index passing rate (within 20% isodose; >98% for 3%/3mm and >92% for 2%/2mm in MapCheck2, and >91% for 3%/3mm criteria in EBT3 film except one case which was for large target and highly modulated). No significant difference (student t-test: p-value < 0.0005) was shown between the doses delivered with and without gating. There was no indication of radiation gap or overlapping during deliver interruption in film dosimetry. Conclusion: The Response™ gating system can be safely used during VMAT treatment. The accurate performance of the gating system linked to ABC can contribute to ITV reduction for SBRT using VMAT.« less

  9. Elevation (typical EBT Coal Hopper Car) with scale. 3 Bay ...

    Library of Congress Historic Buildings Survey, Historic Engineering Record, Historic Landscapes Survey

    Elevation (typical EBT Coal Hopper Car) with scale. 3 Bay Steel Hopper Car with side extensions raising coal carrying capacity to 80,000 pounds. Note cars on either side lack extensions limiting their coal capacity to 70,000 pounds - East Broad Top Railroad & Coal Company, State Route 994, West of U.S. Route 522, Rockhill Furnace, Huntingdon County, PA

  10. Three-dimensional dosimetry of small megavoltage radiation fields using radiochromic gels and optical CT scanning

    NASA Astrophysics Data System (ADS)

    Babic, Steven; McNiven, Andrea; Battista, Jerry; Jordan, Kevin

    2009-04-01

    The dosimetry of small fields as used in stereotactic radiotherapy, radiosurgery and intensity-modulated radiation therapy can be challenging and inaccurate due to partial volume averaging effects and possible disruption of charged particle equilibrium. Consequently, there exists a need for an integrating, tissue equivalent dosimeter with high spatial resolution to avoid perturbing the radiation beam and artificially broadening the measured beam penumbra. In this work, radiochromic ferrous xylenol-orange (FX) and leuco crystal violet (LCV) micelle gels were used to measure relative dose factors (RDFs), percent depth dose profiles and relative lateral beam profiles of 6 MV x-ray pencil beams of diameter 28.1, 9.8 and 4.9 mm. The pencil beams were produced via stereotactic collimators mounted on a Varian 2100 EX linear accelerator. The gels were read using optical computed tomography (CT). Data sets were compared quantitatively with dosimetric measurements made with radiographic (Kodak EDR2) and radiochromic (GAFChromic® EBT) film, respectively. Using a fast cone-beam optical CT scanner (Vista™), corrections for diffusion in the FX gel data yielded RDFs that were comparable to those obtained by minimally diffusing LCV gels. Considering EBT film-measured RDF data as reference, cone-beam CT-scanned LCV gel data, corrected for scattered stray light, were found to be in agreement within 0.5% and -0.6% for the 9.8 and 4.9 mm diameter fields, respectively. The validity of the scattered stray light correction was confirmed by general agreement with RDF data obtained from the same LCV gel read out with a laser CT scanner that is less prone to the acceptance of scattered stray light. Percent depth dose profiles and lateral beam profiles were found to agree within experimental error for the FX gel (corrected for diffusion), LCV gel (corrected for scattered stray light), and EBT and EDR2 films. The results from this study reveal that a three-dimensional dosimetry method

  11. Radiation measurement in the environment of FLASH using passive dosimeters

    NASA Astrophysics Data System (ADS)

    Mukherjee, B.; Rybka, D.; Makowski, D.; Lipka, T.; Simrock, S.

    2007-08-01

    Sophisticated electronic devices comprising sensitive microelectronic components have been installed in the close proximity of the 720 MeV superconducting electron linear accelerator (linac) driving the FLASH (Free Electron Laser in Hamburg), presently in operation at DESY in Hamburg. Microelectronic chips are inherently vulnerable to ionizing radiation, usually generated during routine operation of high-energy particle accelerator facilities like the FLASH. Hence, in order to assess the radiation effect on microelectronic chips and to develop suitable mitigation strategy, it becomes imperative to characterize the radiation field in the FLASH environment. We have evaluated the neutron and gamma energy (spectra) and dose distributions at critical locations in the FLASH tunnel using superheated emulsion (bubble) detectors, GaAs light emitting diodes (LED), LiF-thermoluminescence dosimeters (TLD) and radiochromic (Gafchromic EBT) films. This paper highlights the application of passive dosimeters for an accurate analysis of the radiation field produced by high-energy electron linear accelerators.

  12. Contact radiotherapy using a 50 kV X-ray system: Evaluation of relative dose distribution with the Monte Carlo code PENELOPE and comparison with measurements

    NASA Astrophysics Data System (ADS)

    Croce, Olivier; Hachem, Sabet; Franchisseur, Eric; Marcié, Serge; Gérard, Jean-Pierre; Bordy, Jean-Marc

    2012-06-01

    This paper presents a dosimetric study concerning the system named "Papillon 50" used in the department of radiotherapy of the Centre Antoine-Lacassagne, Nice, France. The machine provides a 50 kVp X-ray beam, currently used to treat rectal cancers. The system can be mounted with various applicators of different diameters or shapes. These applicators can be fixed over the main rod tube of the unit in order to deliver the prescribed absorbed dose into the tumor with an optimal distribution. We have analyzed depth dose curves and dose profiles for the naked tube and for a set of three applicators. Dose measurements were made with an ionization chamber (PTW type 23342) and Gafchromic films (EBT2). We have also compared the measurements with simulations performed using the Monte Carlo code PENELOPE. Simulations were performed with a detailed geometrical description of the experimental setup and with enough statistics. Results of simulations are made in accordance with experimental measurements and provide an accurate evaluation of the dose delivered. The depths of the 50% isodose in water for the various applicators are 4.0, 6.0, 6.6 and 7.1 mm. The Monte Carlo PENELOPE simulations are in accordance with the measurements for a 50 kV X-ray system. Simulations are able to confirm the measurements provided by Gafchromic films or ionization chambers. Results also demonstrate that Monte Carlo simulations could be helpful to validate the future applicators designed for other localizations such as breast or skin cancers. Furthermore, Monte Carlo simulations could be a reliable alternative for a rapid evaluation of the dose delivered by such a system that uses multiple designs of applicators.

  13. EBT reactor systems analysis and cost code: description and users guide (Version 1)

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Santoro, R.T.; Uckan, N.A.; Barnes, J.M.

    1984-06-01

    An ELMO Bumpy Torus (EBT) reactor systems analysis and cost code that incorporates the most recent advances in EBT physics has been written. The code determines a set of reactors that fall within an allowed operating window determined from the coupling of ring and core plasma properties and the self-consistent treatment of the coupled ring-core stability and power balance requirements. The essential elements of the systems analysis and cost code are described, along with the calculational sequences leading to the specification of the reactor options and their associated costs. The input parameters, the constraints imposed upon them, and the operatingmore » range over which the code provides valid results are discussed. A sample problem and the interpretation of the results are also presented.« less

  14. SU-E-T-44: Angular Dependence of Surface Dose Enhancement Measured On Several Inhomogeneities Using Radiochromic EBT3 Films

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jansen, A; Schoenfeld, A; Poppinga, D

    Purpose: The quantification of the relative surface dose enhancement in dependence on the angle of incidence and the atomic number Z of the surface material. Methods: Experiments were performed with slabs made of aluminum, titanium, copper, silver, dental gold and lead. The metal slabs with equal sizes of 1.0×8.0×8.8mm{sup 3} were embedded in an Octavius 4D phantom (PTW Freiburg, Germany). Radiochromic EBT3 films were used to measure the surface dose for angles of incidence ranging from 0° to 90°. The setup with the metals slabs at the isocenter was irradiated with acceleration voltages of 6MV and 10MV. Water reference measurementsmore » were taken under equal conditions. Results: The surface dose enhancement is highest for angles of incidence below 30° and drops significantly for higher. The surface dose enhancement produced by lead and dental gold at 6MV showed a peak of 65%. At 90°, the surface dose enhancement dropped to 15% for both materials. The surface dose enhancements for silver, copper, titanium and aluminum were 45%, 32%, 22% and 12% at 0°, respectively. At an angle of incidence of 80°, the values dropped to 22%, 18%, 12% und 6%. The values for 10MV were very similar. Lead and dental gold showed peaks of 65% und 60%. Their values dropped to 18% at an angle of 90°. The surface dose enhancements for silver, copper, titanium and aluminum were 45%, 30%, 20% and 8% at 0°. At 80° the values dropped to 30%, 20%, 12% and 5%. A dependence of the magnitude of the surface dose enhancement on the atomic number of the surface material can be seen, which is in consistence with literature. Conclusion: The results show that the surface dose enhancements near implant materials with high Z-values should be taken into consideration in radio therapy, even when the angle of incidence is flat.« less

  15. On the suitability of ultrathin detectors for absorbed dose assessment in the presence of high-density heterogeneities.

    PubMed

    Bueno, M; Carrasco, P; Jornet, N; Muñoz-Montplet, C; Duch, M A

    2014-08-01

    The aim of this study was to evaluate the suitability of several detectors for the determination of absorbed dose in bone. Three types of ultrathin LiF-based thermoluminescent dosimeters (TLDs)-two LiF:Mg,Cu,P-based (MCP-Ns and TLD-2000F) and a (7)Li-enriched LiF:Mg,Ti-based (MTS-7s)-as well as EBT2 Gafchromic films were used to measure percentage depth-dose distributions (PDDs) in a water-equivalent phantom with a bone-equivalent heterogeneity for 6 and 18 MV and a set of field sizes ranging from 5 x 5 cm2 to 20 x 20 cm2. MCP-Ns, TLD-2000F, MTS-7s, and EBT2 have active layers of 50, 20, 50, and 30 μm, respectively. Monte Carlo (MC) dose calculations (PENELOPE code) were used as the reference and helped to understand the experimental results and to evaluate the potential perturbation of the fluence in bone caused by the presence of the detectors. The energy dependence and linearity of the TLDs' response was evaluated. TLDs exhibited flat energy responses (within 2.5%) and linearity with dose (within 1.1%) within the range of interest for the selected beams. The results revealed that all considered detectors perturb the electron fluence with respect to the energy inside the bone-equivalent material. MCP-Ns and MTS-7s underestimated the absorbed dose in bone by 4%-5%. EBT2 exhibited comparable accuracy to MTS-7s and MCP-Ns. TLD-2000F was able to determine the dose within 2% accuracy. No dependence on the beam energy or field size was observed. The MC calculations showed that a[Formula: see text] thick detector can provide reliable dose estimations in bone regardless of whether it is made of LiF, water or EBT's active layer material. TLD-2000F was found to be suitable for providing reliable absorbed dose measurements in the presence of bone for high-energy x-ray beams.

  16. TU-F-BRE-08: Significant Variations in Measured Small Cone Output Factor for FFF Beams

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sudhyadhom, A; Ma, L; Kirby, N

    2014-06-15

    Purpose: To evaluate the measurement accuracy of several dosimeters for small cone output factors in two SRS/SBRT dedicated systems with Flattening Filter Free (FFF) beams: a Varian TrueBeam STx (TB) and an Accuray CyberKnife VSI (CK). Output factors (OFs) were measured for both machines and for CK, compared against a Monte Carlo model. Methods: Dose measurements were taken using three different FFF beams (TB 6XFFF, TB 10XFFF, and CK 6XFFF). Three commonly used types of dosimeters were examined in this work: a micro-ion chamber (Exradin A16), two shielded diodes (PTW TN60008 and PTW TN60017), and radiochromic film (Gafchromic EBT2). Measuredmore » OFs from these dosimeters were compared with each other and OFs measured with an Exradin W1 scintillator. Monte Carlo determined correction factors for the CK beam for the micro-ion chamber and diodes were applied to the respective OF measurements and compared against scintillator measured OFs corrected for volume averaging. Results: OFs measured for the smallest fields using the micro-ion chamber, diodes, scintillator, and film varied substantially (with up to a 16% difference between dosimeters). Micro-ion chamber and film OF measurements were up to 9% and 10%, respectively, lower than scintillator measurements for the smallest fields. OF measurements by diode were up to 6% greater than scintillator measurements for the smallest fields. With correction factors, the micro-ion chamber and diode measured OFs showed good agreement with scintillator measured OFs for the CK 6XFFF beam (within 3% and 1.5%, respectively). Conclusion: Uncorrected small field OFs vary significantly with dosimeter. The accuracy of scintillator measurements for small field OFs may be greater than the other dosimeters studied in this work (when uncorrected). Measurements involving EBT2 film may Result in lower accuracy for smaller fields (less than 10mm). Care should be taken in the choice of the dosimeter used for small field OF measurements.« less

  17. Poster — Thur Eve — 30: 4D VMAT dose calculation methodology to investigate the interplay effect: experimental validation using TrueBeam Developer Mode and Gafchromic film

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Teke, T; Milette, MP; Huang, V

    2014-08-15

    The interplay effect between the tumor motion and the radiation beam modulation during a VMAT treatment delivery alters the delivered dose distribution from the planned one. This work present and validate a method to accurately calculate the dose distribution in 4D taking into account the tumor motion, the field modulation and the treatment starting phase. A QUASAR™ respiratory motion phantom was 4D scanned with motion amplitude of 3 cm and with a 3 second period. A static scan was also acquired with the lung insert and the tumor contained in it centered. A VMAT plan with a 6XFFF beam wasmore » created on the averaged CT and delivered on a Varian TrueBeam and the trajectory log file was saved. From the trajectory log file 10 VMAT plans (one for each breathing phase) and a developer mode XML file were created. For the 10 VMAT plans, the tumor motion was modeled by moving the isocentre on the static scan, the plans were re-calculated and summed in the treatment planning system. In the developer mode, the tumor motion was simulated by moving the couch dynamically during the treatment. Gafchromic films were placed in the QUASAR phantom static and irradiated using the developer mode. Different treatment starting phase were investigated (no phase shift, maximum inhalation and maximum exhalation). Calculated and measured isodose lines and profiles are in very good agreement. For each starting phase, the dose distribution exhibit significant differences but are accurately calculated with the methodology presented in this work.« less

  18. Quantitative assessment of the accuracy of dose calculation using pencil beam and Monte Carlo algorithms and requirements for clinical quality assurance

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ali, Imad, E-mail: iali@ouhsc.edu; Ahmad, Salahuddin

    2013-10-01

    To compare the doses calculated using the BrainLAB pencil beam (PB) and Monte Carlo (MC) algorithms for tumors located in various sites including the lung and evaluate quality assurance procedures required for the verification of the accuracy of dose calculation. The dose-calculation accuracy of PB and MC was also assessed quantitatively with measurement using ionization chamber and Gafchromic films placed in solid water and heterogeneous phantoms. The dose was calculated using PB convolution and MC algorithms in the iPlan treatment planning system from BrainLAB. The dose calculation was performed on the patient's computed tomography images with lesions in various treatmentmore » sites including 5 lungs, 5 prostates, 4 brains, 2 head and necks, and 2 paraspinal tissues. A combination of conventional, conformal, and intensity-modulated radiation therapy plans was used in dose calculation. The leaf sequence from intensity-modulated radiation therapy plans or beam shapes from conformal plans and monitor units and other planning parameters calculated by the PB were identical for calculating dose with MC. Heterogeneity correction was considered in both PB and MC dose calculations. Dose-volume parameters such as V95 (volume covered by 95% of prescription dose), dose distributions, and gamma analysis were used to evaluate the calculated dose by PB and MC. The measured doses by ionization chamber and EBT GAFCHROMIC film in solid water and heterogeneous phantoms were used to quantitatively asses the accuracy of dose calculated by PB and MC. The dose-volume histograms and dose distributions calculated by PB and MC in the brain, prostate, paraspinal, and head and neck were in good agreement with one another (within 5%) and provided acceptable planning target volume coverage. However, dose distributions of the patients with lung cancer had large discrepancies. For a plan optimized with PB, the dose coverage was shown as clinically acceptable, whereas in reality, the MC

  19. Use of a control film piece in radiochromic film dosimetry.

    PubMed

    Aldelaijan, Saad; Alzorkany, Faisal; Moftah, Belal; Buzurovic, Ivan; Seuntjens, Jan; Tomic, Nada; Devic, Slobodan

    2016-01-01

    Radiochromic films change their color upon irradiation due to polymerization of the sensitive component embedded within the sensitive layer. However, agents, other than monitored radiation, can lead to a change in the color of the sensitive layer (temperature, humidity, UV light) that can be considered as a background signal and can be removed from the actual measurement by using a control film piece. In this work, we investigate the impact of the use of control film pieces on both accuracy and uncertainty of dose measured using radiochromic film based reference dosimetry protocol. We irradiated "control" film pieces (EBT3 GafChromic(TM) film model) to known doses in a range of 0.05-1 Gy, and five film pieces of the same size to 2, 5, 10, 15 and 20 Gy, considered to be "unknown" doses. Depending on a dose range, two approaches to incorporating control film piece were investigated: signal and dose corrected method. For dose values greater than 10 Gy, the increase in accuracy of 3% led to uncertainty loss of 5% by using dose corrected approach. At lower doses and signals of the order of 5%, we observed an increase in accuracy of 10% with a loss of uncertainty lower than 1% by using the corrected signal approach. Incorporation of the signal registered by the control film piece into dose measurement analysis should be a judgment call of the user based on a tradeoff between deemed accuracy and acceptable uncertainty for a given dose measurement. Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  20. Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE®/ optical-CT 3D dosimetry system

    NASA Astrophysics Data System (ADS)

    Clift, Corey; Thomas, Andrew; Adamovics, John; Chang, Zheng; Das, Indra; Oldham, Mark

    2010-03-01

    Achieving accurate small field dosimetry is challenging. This study investigates the utility of a radiochromic plastic PRESAGE® read with optical-CT for the acquisition of radiosurgery field commissioning data from a Novalis Tx system with a high-definition multileaf collimator (HDMLC). Total scatter factors (Sc, p), beam profiles, and penumbrae were measured for five different radiosurgery fields (5, 10, 20, 30 and 40 mm) using a commercially available optical-CT scanner (OCTOPUS, MGS Research). The percent depth dose (PDD), beam profile and penumbra of the 10 mm field were also measured using a higher resolution in-house prototype CCD-based scanner. Gafchromic EBT® film was used for independent verification. Measurements of Sc, p made with PRESAGE® and film agreed with mini-ion chamber commissioning data to within 4% for every field (range 0.2-3.6% for PRESAGE®, and 1.6-3.6% for EBT). PDD, beam profile and penumbra measurements made with the two PRESAGE®/optical-CT systems and film showed good agreement with the high-resolution diode commissioning measurements with a competitive resolution (0.5 mm pixels). The in-house prototype optical-CT scanner allowed much finer resolution compared with previous applications of PRESAGE®. The advantages of the PRESAGE® system for small field dosimetry include 3D measurements, negligible volume averaging, directional insensitivity, an absence of beam perturbations, energy and dose rate independence.

  1. Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE®/optical-CT 3D dosimetry system

    PubMed Central

    Clift, Corey; Thomas, Andrew; Adamovics, John; Chang, Zheng; Das, Indra; Oldham, Mark

    2010-01-01

    Achieving accurate small field dosimetry is challenging. This study investigates the utility of a radiochromic plastic PRESAGE® read with optical-CT for the acquisition of radiosurgery field commissioning data from a Novalis Tx system with a high-definition multileaf collimator (HDMLC). Total scatter factors (Sc, p), beam profiles, and penumbrae were measured for five different radiosurgery fields (5, 10, 20, 30 and 40 mm) using a commercially available optical-CT scanner (OCTOPUS, MGS Research). The percent depth dose (PDD), beam profile and penumbra of the 10 mm field were also measured using a higher resolution in-house prototype CCD-based scanner. Gafchromic EBT® film was used for independent verification. Measurements of Sc, p made with PRESAGE® and film agreed with mini-ion chamber commissioning data to within 4% for every field (range 0.2–3.6% for PRESAGE®, and 1.6–3.6% for EBT). PDD, beam profile and penumbra measurements made with the two PRESAGE®/optical-CT systems and film showed good agreement with the high-resolution diode commissioning measurements with a competitive resolution (0.5 mm pixels). The in-house prototype optical-CT scanner allowed much finer resolution compared with previous applications of PRESAGE®. The advantages of the PRESAGE® system for small field dosimetry include 3D measurements, negligible volume averaging, directional insensitivity, an absence of beam perturbations, energy and dose rate independence. PMID:20134082

  2. SU-G-201-05: Comparison of Different Methods for Output Verification of Eleckta Nucletron’s Valencia Skin Applicators

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Barrett, J; Yudelev, M

    2016-06-15

    Purpose: The provided output factors for Elekta Nucletron’s skin applicators are based on Monte Carlo simulations. These outputs have not been independently verified, and there is no recognized method for output verification of the vendor’s applicators. The purpose of this work is to validate the outputs provided by the vendor experimentally. Methods: Using a Flexitron Ir-192 HDR unit, three experimental methods were employed to determine dose with the 30 mm diameter Valencia applicator: first a gradient method using extrapolation ionization chamber (Far West Technology, EIC-1) measurements in solid water phantom at 3 mm SCD was used. The dose was derivedmore » based on first principles. Secondly a combination of a parallel plate chamber (Exradin A-10) and the EIC-1 was used to determine air kerma at 3 mm SCD. The air kerma was converted to dose to water in line with TG-61 formalism by using a muen ratio and a scatter factor measured with the skin applicators. Similarly a combination of the A-10 parallel plate chamber and gafchromic film (EBT 3) was also used. The Nk factor for the A-10 chamber was obtained through linear interpolation between ADCL supplied Nk factors for Cs-137 and M250. Results: EIC-1 measurements in solid water defined the outputs factor at 3 mm as 0.1343 cGy/U hr. The combination of A-10/ EIC-1 and A-10/EBT3 lead to output factors of 0.1383 and 0.1568 cGy/U hr, respectively. For comparison the output recommended by the vendor is 0.1659 cGy/U hr. Conclusion: All determined dose rates were lower than the vendor supplied values. The observed discrepancy between extrapolation chamber and film methods can be ascribed to extracameral gradient effects that may not be fully accounted for by the former method.« less

  3. Poster — Thur Eve — 58: Dosimetric validation of electronic compensation for radiotherapy treatment planning

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Gräfe, James; Khan, Rao; Meyer, Tyler

    2014-08-15

    In this study we investigate the deliverability of dosimetric plans generated by the irregular surface compensator (ISCOMP) algorithm for 6 MV photon beams in Eclipse (Varian Medical System, CA). In contrast to physical tissue compensation, the electronic ISCOMP uses MLCs to dynamically modulate the fluence of a photon beam in order to deliver a uniform dose at a user defined plane in tissue. This method can be used to shield critical organs that are located within the treatment portal or improve dose uniformity by tissue compensation in inhomogeneous regions. Three site specific plans and a set of test fields weremore » evaluated using the γ-metric of 3%/ 3 mm on Varian EPID, MapCHECK, and Gafchromic EBT3 film with a clinical tolerance of >95% passing rates. Point dose measurements with an NRCC calibrated ionization chamber were also performed to verify the absolute dose delivered. In all cases the MapCHECK measured plans met the gamma criteria. The mean passing rate for the six EBT3 film field measurements was 96.2%, with only two fields at 93.4 and 94.0% passing rates. The EPID plans passed for fields encompassing the central ∼10 × 10 cm{sup 2} region of the detector; however for larger fields and greater off-axis distances discrepancies were observed and attributed to the profile corrections and modeling of backscatter in the portal dose calculation. The magnitude of the average percentage difference for 21 ion chamber point dose measurements and 17 different fields was 1.4 ± 0.9%, and the maximum percentage difference was −3.3%. These measurements qualify the algorithm for routine clinical use subject to the same pre-treatment patient specific QA as IMRT.« less

  4. Comparison of selected dose calculation algorithms in radiotherapy treatment planning for tissues with inhomogeneities

    NASA Astrophysics Data System (ADS)

    Woon, Y. L.; Heng, S. P.; Wong, J. H. D.; Ung, N. M.

    2016-03-01

    Inhomogeneity correction is recommended for accurate dose calculation in radiotherapy treatment planning since human body are highly inhomogeneous with the presence of bones and air cavities. However, each dose calculation algorithm has its own limitations. This study is to assess the accuracy of five algorithms that are currently implemented for treatment planning, including pencil beam convolution (PBC), superposition (SP), anisotropic analytical algorithm (AAA), Monte Carlo (MC) and Acuros XB (AXB). The calculated dose was compared with the measured dose using radiochromic film (Gafchromic EBT2) in inhomogeneous phantoms. In addition, the dosimetric impact of different algorithms on intensity modulated radiotherapy (IMRT) was studied for head and neck region. MC had the best agreement with the measured percentage depth dose (PDD) within the inhomogeneous region. This was followed by AXB, AAA, SP and PBC. For IMRT planning, MC algorithm is recommended for treatment planning in preference to PBC and SP. The MC and AXB algorithms were found to have better accuracy in terms of inhomogeneity correction and should be used for tumour volume within the proximity of inhomogeneous structures.

  5. Surface dosimetry for breast radiotherapy in the presence of immobilization cast material

    NASA Astrophysics Data System (ADS)

    Kelly, Andrew; Hardcastle, Nicholas; Metcalfe, Peter; Cutajar, Dean; Quinn, Alexandra; Foo, Kerwyn; Cardoso, Michael; Barlin, Sheree; Rosenfeld, Anatoly

    2011-02-01

    Curative breast radiotherapy typically leaves patients with varying degrees of cosmetic damage. One problem interfering with cosmetically acceptable breast radiotherapy is the external contour for large pendulous breasts which often results in high doses to skin folds. Thermoplastic casts are often employed to secure the breasts to maintain setup reproducibility and limit the presence of skin folds. This paper aims to determine changes in surface dose that can be attributed to the use of thermoplastic immobilization casts. Skin dose for a clinical hybrid conformal/IMRT breast plan was measured using radiochromic film and MOSFET detectors at a range of water equivalent depths representative of the different skin layers. The radiochromic film was used as an integrating dosimeter, while the MOSFETs were used for real-time dosimetry to isolate the contribution of skin dose from individual IMRT segments. Strips of film were placed at various locations on the breast and the MOSFETs were used to measure skin dose at 16 positions spaced along the film strips for comparison of data. The results showed an increase in skin dose in the presence of the immobilization cast of up to 45.7% and 62.3% of the skin dose without the immobilization cast present as measured with Gafchromic EBT film and MOSFETs, respectively. The increase in skin dose due to the immobilization cast varied with the angle of beam incidence and was greatest when the beam was normally incident on the phantom. The increase in surface dose with the immobilization cast was greater under entrance dose conditions compared to exit dose conditions.

  6. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Escarcia, F; Herrera, J; Garcia, O

    Purpose: To investigate the uniformity of film response of EBT3 film of two different film batches. It has been reported that the response of the EBT film family is not homogenous between film batches. The later may have an impact in the dosimetry of small radiotherapy beams. Methods: A solid water phantom was used for dosimetric measurements. EBT3 film irradiation was performed with a 6 MV photon beam at 5 cm depth with a SAD of 100 cm. All irradiations were performed perpendicularly to the film plane covering the dose range 1 to 10 Gy. Three square field sizes weremore » used to analyze the film response energy dependence: 10, 5 and 1 cm{sup 2}. Two batches of film EBT3 were used #A03181302 (B1) and #03031403 (B2). Film read out was carrying out with an Epson Perfection V750-Pro flatbed scanner in transmission mode with a spatial resolution of 72 dpi, with all post-processing and colour management options turned off, using 48 bits RGB colour depth. The scans were analyzed with the red channel. Results: The results shown that there were differences between the film response for each batch. The differences between batches for 1 Gy were 2%, 6% and 12% for 10, 5 and 1 cm2 square field sizes, respectively. The differences found for 10 Gy were 13%, 14% and 13% for 10, 5 and 1 cm{sup 2} square field sizes, respectively. It can be observed that the dependence with field size dismissed for higher doses. The later may be due to film response saturation at 10 Gy. Conclusion: The EBT3 film -as its predecessors-, it suffer for inter-batch variability in the film response. Further research is required to assess the possible impact in small beam dosimetry.« less

  7. Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)

    NASA Astrophysics Data System (ADS)

    Purwaningsih, S.; Lubis, L. E.; Pawiro, S. A.; Soejoko, D. S.

    2016-03-01

    This research was aimed to check the patterns of dose profile on adult and pediatric head scan. We compared measurement result on dose profile along the z- axis rotation at peripheries and center phantom with a variety of pitch, i.e. 0.75, 1, 1.5 for adult and pediatric head protocol, keeping the rest of the scan parameters constant. Measurements were performed on homogeneous, cylindrical PMMA phantom with diameters of 16 and 10 cm using XR-QA2 Gafchromic film and TLD as dosimeters. The measurement result indicated a decrease in the dose about 50% and 47% for adult and pediatric head scan with the increase of pitch. For 0.75 value of pitch adult head scan, dose range for each position were (2.4 - 5.0) cGy, (3.1 - 5.3) cGy, (2.2 - 4.5) cGy, (2.8 - 5.3) cGy, and (3.3 - 5.6) cGy for position of center, 3, 6, 9 and 12 o'clock peripheral phantom position respectively. Dose profile for adult and pediatric head scan protocols has pattern curve with the maximum dose in the middle and tendency of symmetry near the edges, with different the plateau length along z- axis direction in accordance to the measurement position in the phantom.

  8. SU-E-T-506: Intercomparison Study On Small Field Output Factor Measurements

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Talamonti, C; Casati, M; Compagnucci, A

    2015-06-15

    Purpose In radiotherapy, uncertainties due to small field measurements (SFM) introduce systematic errors to the treatment process and the development of new dosimeters for quality assurance programs is a challenge. In this work we analyze the behavior of seven detectors measuring output factors of 6MV photon beam. Methods The dosimeters employed are: a single cristal diamond detector (SCCD) developed at the University of Rome Tor Vergata, a silicon diode developed within the project MAESTRO, a IBA Razor silicon diode, A1SL and A26 Exradin ion chambers, an EBT3 Gafchromic film and the Exradin W1 Scintillator.Diamond sensitive volume is a cylinder 2.2mmmore » in diameter and 1μm thick. MAESTRO diode is 2×2mm2 active area. Razor sensitive volume is a cylinder 0.6 mm in diameter and 0.02 mm thick. A16 and A1Sl have a collecting volume of 0,015cc and 0,053cc. The W1 is an optical fiber with an active volume of 0.002cc. All measurements were performed in a water phantom, with detector positioned at the isocenter (SSD=90cm, d=10cm), MAESTRO diode being sandwiched in solid water to obtain an equivalent experimental setup. Results These measurements are challenging due to the absence of charged particle equilibrium conditions, detector size and positioning problems. They are in good agreement among each other, especially GAF, Razor, W1 and SCDD. Maximum deviations reported are related to the field 0.8×0.8cm2 for MAESTRO and chambers data with respect to EBT3: around 15% (A1SLvsEBT3), 16% (MAESTROvsEBT3). Razor and W1 show a deviation around 3% with respect to SCDD. Conclusion In this work measurements made with a variety of detectors are compared. These study show the possibility to choose different detectors for SFM and that smaller ion chambers are still not competitive with solid state detectors. Silicon, diamond and optical fiber dosimeters show a similar behavior with minor discrepancies for the smallest field.« less

  9. Experimental assessment of the Advanced Collapsed-cone Engine for scalp brachytherapy treatments.

    PubMed

    Cawston-Grant, Brie; Morrison, Hali; Sloboda, Ron S; Menon, Geetha

    To experimentally assess the performance of the Advanced Collapsed-cone Engine (ACE) for 192 Ir high-dose-rate brachytherapy treatment planning of nonmelanoma skin cancers of the scalp. A layered slab phantom was designed to model the head (skin, skull, and brain) and surface treatment mold using tissue equivalent materials. Six variations of the phantom were created by varying skin thickness, skull thickness, and size of air gap between the mold and skin. Treatment planning was initially performed using the Task Group 43 (TG-43) formalism with CT images of each phantom variation. Doses were recalculated using standard and high accuracy modes of ACE. The plans were delivered to Gafchromic EBT3 film placed between different layers of the phantom. Doses calculated by TG-43 and ACE and those measured by film agreed with each other at most locations within the phantoms. For a given phantom variation, average TG-43- and ACE-calculated doses were similar, with a maximum difference of (3 ± 12)% (k = 2). Compared to the film measurements, TG-43 and ACE overestimated the film-measured dose by (13 ± 12)% (k = 2) for one phantom variation below the skull layer. TG-43- and ACE-calculated and film-measured doses were found to agree above the skull layer of the phantom, which is where the tumor would be located in a clinical case. ACE appears to underestimate the attenuation through bone relative to that measured by film; however, the dose to bone is below tolerance levels for this treatment. Copyright © 2017 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  10. Dosimetry of Strontium eye applicator: Comparison of Monte Carlo calculations and radiochromic film measurements

    NASA Astrophysics Data System (ADS)

    Laoues, M.; Khelifi, R.; Moussa, A. S.

    2015-01-01

    Strontium-90 eye applicators are a beta-ray emitter with a relatively high-energy (maximum energy about 2.28 MeV and average energy about 0.9 MeV). These applicators come in different shapes and dimensions; they are used for the treatment of eye diseases. Whenever, radiation is used in treatment, dosimetry is essential. However, knowledge of the exact dose distribution is a critical decision-making to the outcome of the treatment. The main aim of our study is to simulate the dosimetry of the SIA.20 eye applicator with Monte Carlo GATE 6.1 platform and to compare the calculated results with those measured with EBT2 films. This means that GATE and EBT2 were used to quantify the surface and depths dose- rate, the relative dose profile and the dosimetric parameters in according to international recommendations. Calculated and measured results are in good agreement and they are consistent with the ICRU and NCS recommendations.

  11. Radiochromic film sensitivity calibrations using ion beams from a Pelletron accelerator

    NASA Astrophysics Data System (ADS)

    Filkins, T. M.; Steidle, Jessica; Ward, R. J.; Freeman, C. G.; Padalino, S. J.; Regan, S. P.; Sangster, T. C.

    2015-11-01

    Radiochromic film (RCF) is a transparent detector film that permanently changes color following exposure to ionizing radiation. The optical density of the film increases with increasing absorbed dose. RCF is convenient to use because it requires no chemical processing and can be scanned using commercially available document scanners. RCF is used frequently in medical applications, but is also used in a variety of diagnostics in high energy density physics. The film consists of a single or double layer of radiation-sensitive organic microcrystal monomers placed onto a polyester backing. GafchromicTM manufactures a large number of different types of RCF, and new types of film frequently replace older products. In this study, the sensitivity of several types of RCF to ion beams of different energies was measured. Ion beams produced by the SUNY Geneseo 1.7 MV Pelletron accelerator were directed into a target chamber where they scattered off of a gold foil. A sample of RCF was exposed to the scattered ions. The fluence of incident particles on the film was measured using a surface barrier detector. Results of these calibrations will be presented. This work was funded in part by a grant from the DOE through the Laboratory for Laser Energetics.

  12. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rosen, B. S., E-mail: bsrosen@wisc.edu; Hammer, C. G.; Kunugi, K. A.

    Purpose: To evaluate a prototype densitometer traceable to primary optical standards and compare its performance to an EPSON Expression{sup ®} 10000XL flatbed scanner (the Epson) for quantitative radiochromic film (RCF) dosimetry. Methods: A prototype traceable laser densitometry system (LDS) was developed to mitigate common film scanning artifacts, such as positional scan dependence and high noise in low-dose regions, by performing point-based measurements of RCF suspended in free-space using coherent light. The LDS and the Epson optical absorbance scales were calibrated up to 3 AU, using reference materials calibrated at a primary standards laboratory and a scanner calibration factor (SCF). Calibratedmore » optical density (OD) was determined for 96 Gafchromic{sup ®} EBT3 film segments before and after irradiation to one of 16 dose levels between 0 and 10 Gy, exposed to {sup 60}Co in a polymethyl-methacrylate (PMMA) phantom. The sensitivity was determined at each dose level and at two rotationally orthogonal readout orientations to obtain the sensitometric response of each RCF dosimetry system. LDS rotational scanning dependence was measured at nine angles between 0°and 180°, due to the expected interference between coherent light and polarizing EBT3 material. The response curves were fit to the analytic functions predicted by two physical response models: the two-parameter single-hit model and the four-parameter percolation model. Results: The LDS and the Epson absorbance measurements were linear to primary optical standards to within 0.2% and 0.3% up to 2 and 1 AU, respectively. At higher densities, the LDS had an over-response (2.5% at 3 AU) and the Epson an under-response (3.1% and 9.8% at 2 and 3 AU, respectively). The LDS and the Epson SCF over the applicable range were 0.968% ± 0.2% and 1.561% ± 0.3%, respectively. The positional scan dependence was evaluated on each digitizer and shown to be mitigated on the LDS, as compared to the Epson. Maximum

  13. SU-F-T-70: A High Dose Rate Total Skin Electron Irradiation Technique with A Specific Inter-Film Variation Correction Method for Very Large Electron Beam Fields

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Yang, X; Rosenfield, J; Dong, X

    2016-06-15

    Purpose: Rotational total skin electron irradiation (RTSEI) is used in the treatment of cutaneous T-cell lymphoma. Due to inter-film uniformity variations the dosimetry measurement of a large electron beam of a very low energy is challenging. This work provides a method to improve the accuracy of flatness and symmetry for a very large treatment field of low electron energy used in dual beam RTSEI. Methods: RTSEI is delivered by dual angles field a gantry of ±20 degrees of 270 to cover the upper and the lower halves of the patient body with acceptable beam uniformity. The field size is inmore » the order of 230cm in vertical height and 120 cm in horizontal width and beam energy is a degraded 6 MeV (6 mm of PMMA spoiler). We utilized parallel plate chambers, Gafchromic films and OSLDs as a measuring devices for absolute dose, B-Factor, stationary and rotational percent depth dose and beam uniformity. To reduce inter-film dosimetric variation we introduced a new specific correction method to analyze beam uniformity. This correction method uses some image processing techniques combining film value before and after radiation dose to compensate the inter-variation dose response differences among films. Results: Stationary and rotational depth of dose demonstrated that the Rp is 2 cm for rotational and the maximum dose is shifted toward the surface (3mm). The dosimetry for the phantom showed that dose uniformity reduced to 3.01% for the vertical flatness and 2.35% for horizontal flatness after correction thus achieving better flatness and uniformity. The absolute dose readings of calibrated films after our correction matched with the readings from OSLD. Conclusion: The proposed correction method for Gafchromic films will be a useful tool to correct inter-film dosimetric variation for the future clinical film dosimetry verification in very large fields, allowing the optimizations of other parameters.« less

  14. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Morrison, H; Menon, G; Sloboda, R

    The purpose of this study was to investigate the accuracy of radiochromic film calibration procedures used in external beam radiotherapy when applied to I-125 brachytherapy sources delivering higher doses, and to determine any necessary modifications to achieve similar accuracy in absolute dose measurements. GafChromic EBT3 film was used to measure radiation doses upwards of 35 Gy from 6 MV, 75 kVp and (∼28 keV) I-125 photon sources. A custom phantom was used for the I-125 irradiations to obtain a larger film area with nearly constant dose to reduce the effects of film heterogeneities on the optical density (OD) measurements. RGBmore » transmission images were obtained with an Epson 10000XL flatbed scanner, and calibration curves relating OD and dose using a rational function were determined for each colour channel and at each energy using a non-linear least square minimization method. Differences found between the 6 MV calibration curve and those for the lower energy sources are large enough that 6 MV beams should not be used to calibrate film for low-energy sources. However, differences between the 75 kVp and I-125 calibration curves were quite small; indicating that 75 kVp is a good choice. Compared with I-125 irradiation, this gives the advantages of lower type B uncertainties and markedly reduced irradiation time. To obtain high accuracy calibration for the dose range up to 35 Gy, two-segment piece-wise fitting was required. This yielded absolute dose measurement accuracy above 1 Gy of ∼2% for 75 kVp and ∼5% for I-125 seed exposures.« less

  15. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Schoenfeld, A; Poppinga, D; Poppe, B

    Purpose: This study aims to investigate the optical properties of radiochromic EBT3 films on exposure to polarized incident light. Methods: An optical table setup was used to investigate the properties of exposed and unexposed EBT3 films. The films were placed with their long side horizontally and illuminated with polarized incident white light. The polarization of light with the electrical vector pointing vertically is referred to as 0°, accordingly horizontal orientation corresponds to 90°. The light transmission was measured depending on the polarization angle of the incident light and the polarization of a polarizer in front of the detector. Secondly, themore » scattering properties of exposed and unexposed films were measured by placing a plane convex lens behind the films and a screen in its focal plane. Thereby, the distribution of the scattering angles appears as an intensity map on the screen. The distributions of scattering angles caused by EBT3 films and by neutral density filters were compared. Results: EBT3 films show a strong dependence of the light transmission on the polarization of the incident light. With both polarizers parallel, a peak transmission was found at 90° orientation of the polarizers. With the rear polarizer at right angles with the front polarizer, peak transmissions were found at front polarizer orientations 45° and 135°. The scattering appears to be anisotropic with a preference direction parallel to the long side of the film. The portion of scattered light and the half value scattering angle both increase with the dose on the film. Conclusion: EBT3 films show dose dependent changes in polarized light transmission and anisotropic light scattering. These effects impair the light absorption measurements on exposed films performed with commercial flatbed scanners and are causing the well-known artifacts of radiochromic film dosimetry with flatbed scanners, the “orientation effect” and the “parabola effect”.« less

  16. A respiratory compensating system: design and performance evaluation.

    PubMed

    Chuang, Ho-Chiao; Huang, Ding-Yang; Tien, Der-Chi; Wu, Ren-Hong; Hsu, Chung-Hsien

    2014-05-08

    This study proposes a respiratory compensating system which is mounted on the top of the treatment couch for reverse motion, opposite from the direction of the targets (diaphragm and hemostatic clip), in order to offset organ displacement generated by respiratory motion. Traditionally, in the treatment of cancer patients, doctors must increase the field size for radiation therapy of tumors because organs move with respiratory motion, which causes radiation-induced inflammation on the normal tissues (organ at risk (OAR)) while killing cancer cells, and thereby reducing the patient's quality of life. This study uses a strain gauge as a respiratory signal capture device to obtain abdomen respiratory signals, a proposed respiratory simulation system (RSS) and respiratory compensating system to experiment how to offset the organ displacement caused by respiratory movement and compensation effect. This study verifies the effect of the respiratory compensating system in offsetting the target displacement using two methods. The first method uses linac (medical linear accelerator) to irradiate a 300 cGy dose on the EBT film (GAFCHROMIC EBT film). The second method uses a strain gauge to capture the patients' respiratory signals, while using fluoroscopy to observe in vivo targets, such as a diaphragm, to enable the respiratory compensating system to offset the displacements of targets in superior-inferior (SI) direction. Testing results show that the RSS position error is approximately 0.45 ~ 1.42 mm, while the respiratory compensating system position error is approximately 0.48 ~ 1.42 mm. From the EBT film profiles based on different input to the RSS, the results suggest that when the input respiratory signals of RSS are sine wave signals, the average dose (%) in the target area is improved by 1.4% ~ 24.4%, and improved in the 95% isodose area by 15.3% ~ 76.9% after compensation. If the respiratory signals input into the RSS respiratory signals are actual human respiratory

  17. The treatment of extensive scalp lesions combining electrons with intensity-modulated photons.

    PubMed

    Chan, Maria F; Song, Yulin; Burman, Chandra; Chui, Chen S; Schupak, Karen

    2006-01-01

    This study was to investigate the feasibility and potential benefits of combining electrons with intensity modulated photons (IMRT+e) for patients with extensive scalp lesions. A case of a patient with an extensive scalp lesion, in which the target volume covered the entire front half of the scalp, is presented. This approach incorporated the electron dose into the inverse treatment planning optimization. The resulting doses to the planning target volume (PTV) and relevant critical structures were compared. Thermoluminescent dosimeters (TLD), diodes, and GAFCHROMIC EBT films were used to verify the accuracy of the techniques. The IMRT+e plan produced a superior dose distribution to the patient as compared to the IMRT plan in terms of reduction of the dose to the brain with the same dose conformity and homogeneity in the target volumes. This study showed that IMRT+e is a viable treatment modality for extensive scalp lesions patients. It provides a feasible alternative to existing treatment techniques, resulting in improved homogeneity of dose to the PTV compared to conventional electron techniques and a decrease in dose to the brain compared to photon IMRT alone.

  18. Commissioning a hobby cutting device for radiochromic film preparation.

    PubMed

    Zolfaghari, Somayeh; Francis, Kirby E; Kairn, Tanya; Crowe, Scott B

    2017-06-01

    In addition to a high spatial resolution and well characterised dose response, one of the major advantages of radiochromic film as a dosimeter is that sheets of film can be cut into pieces suitable for use as calibration films, and for in vivo and phantom measurements. The cutting of film is typically done using scissors or a guillotine, and this process can be time-consuming, limited in precision, requires extensive handling and does not allow holes to be cut from the film without cutting from an existing edge. This study investigated the use of a Brother ScanNCut hobby cutting system for EBT3 film preparation. The optimal operating parameters (blade size, pressure, speed) that resulted in precise cuts with minimal delamination at cut edges were identified using test cutting patterns. These parameters were then used to cut a large film insert for a stereotactic head phantom for comparison against an insert cut with scissors. While the hobby cutting system caused a wider region of delamination at the film edge (1.8 mm) compared to scissors (1 mm), the hobby cutting system was found to be able to produce reproducible cuts more efficiently and more accurately than scissors. The use of the hobby cutting system is recommended for complex phantom inserts (containing sharp corners or holes for alignment rods) or in situations where large numbers of film pieces need to be prepared.

  19. Application of a dummy eye shield for electron treatment planning

    PubMed Central

    Kang, Sei-Kwon; Park, Soah; Hwang, Taejin; Cheong, Kwang-Ho; Han, Taejin; Kim, Haeyoung; Lee, Me-Yeon; Kim, Kyoung Ju; Oh, Do Hoon; Bae, Hoonsik

    2013-01-01

    Metallic eye shields have been widely used for near-eye treatments to protect critical regions, but have never been incorporated into treatment plans because of the unwanted appearance of the metal artifacts on CT images. The purpose of this work was to test the use of an acrylic dummy eye shield as a substitute for a metallic eye shield during CT scans. An acrylic dummy shield of the same size as the tungsten eye shield was machined and CT scanned. The BEAMnrc and the DOSXYZnrc were used for the Monte Carlo (MC) simulation, with the appropriate material information and density for the aluminum cover, steel knob and tungsten body of the eye shield. The Pinnacle adopting the Hogstrom electron pencil-beam algorithm was used for the one-port 6-MeV beam plan after delineation and density override of the metallic parts. The results were confirmed with the metal oxide semiconductor field effect transistor (MOSFET) detectors and the Gafchromic EBT2 film measurements. For both the maximum eyelid dose over the shield and the maximum dose under the shield, the MC results agreed with the EBT2 measurements within 1.7%. For the Pinnacle plan, the maximum dose under the shield agreed with the MC within 0.3%; however, the eyelid dose differed by –19.3%. The adoption of the acrylic dummy eye shield was successful for the treatment plan. However, the Pinnacle pencil-beam algorithm was not sufficient to predict the eyelid dose on the tungsten shield, and more accurate algorithms like MC should be considered for a treatment plan. PMID:22915776

  20. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Menegotti, L.; Delana, A.; Martignano, A.

    Film dosimetry is an attractive tool for dose distribution verification in intensity modulated radiotherapy (IMRT). A critical aspect of radiochromic film dosimetry is the scanner used for the readout of the film: the output needs to be calibrated in dose response and corrected for pixel value and spatial dependent nonuniformity caused by light scattering; these procedures can take a long time. A method for a fast and accurate calibration and uniformity correction for radiochromic film dosimetry is presented: a single film exposure is used to do both calibration and correction. Gafchromic EBT films were read with two flatbed charge coupledmore » device scanners (Epson V750 and 1680Pro). The accuracy of the method is investigated with specific dose patterns and an IMRT beam. The comparisons with a two-dimensional array of ionization chambers using a 18x18 cm{sup 2} open field and an inverse pyramid dose pattern show an increment in the percentage of points which pass the gamma analysis (tolerance parameters of 3% and 3 mm), passing from 55% and 64% for the 1680Pro and V750 scanners, respectively, to 94% for both scanners for the 18x18 open field, and from 76% and 75% to 91% for the inverse pyramid pattern. Application to an IMRT beam also shows better gamma index results, passing from 88% and 86% for the two scanners, respectively, to 94% for both. The number of points and dose range considered for correction and calibration appears to be appropriate for use in IMRT verification. The method showed to be fast and to correct properly the nonuniformity and has been adopted for routine clinical IMRT dose verification.« less

  1. Optically stimulated luminescence (OSL) of carbon-doped aluminum oxide (Al{sub 2}O{sub 3}:C) for film dosimetry in radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Schembri, V.; Heijmen, B. J. M.

    2007-06-15

    Introduction and Purpose: Conventional x-ray films and radiochromic films have inherent challenges for high precision radiotherapy dosimetry. Here we have investigated basic characteristics of optically stimulated luminescence (OSL) of irradiated films containing carbon-doped aluminum oxide (Al{sub 2}O{sub 3}:C) for dosimetry in therapeutic photon and electron beams. Materials and Methods: The OSL films consist of a polystyrene sheet, with a top layer of a mixture of single crystals of Al{sub 2}O{sub 3}:C, ground into a powder, and a polyester base. The total thickness of the films is 0.3 mm. Measurements have been performed in a water equivalent phantom, using 4, 6,more » 10, and 18 MV photon beams, and 6-22 MeV electron beams. The studies include assessment of the film response (acquired OSL signal/delivered dose) on delivered dose (linearity), dose rate (1-6 Gy/min), beam quality, field size and depth (6 MV, ranges 4x4-30x30 cm{sup 2}, d{sub max}-35 cm). Doses have been derived from ionization chamber measurements. OSL films have also been compared with conventional x-ray and GafChromic films for dosimetry outside the high dose area, with a high proportion of low dose scattered photons. In total, 787 OSL films have been irradiated. Results: Overall, the OSL response for electron beams was 3.6% lower than for photon beams. Differences between the various electron beam energies were not significant. The 6 and 18 MV photon beams differed in response by 4%. No response dependencies on dose rate were observed. For the 6 MV beam, the field size and depth dependencies of the OSL response were within {+-}2.5%. The observed inter-film response variation for films irradiated with the same dose varied from 1% to 3.2% (1 SD), depending on the measurement day. At a depth of 20 cm, 5 cm outside the 20x20 cm{sup 2} 6 and 18 MV beams, an over response of 17% was observed. In contrast to GafChromic and conventional x-ray films, the response of the Al{sub 2}O{sub 3}:C films is

  2. Evaluation of the dosimetric properties of a diode detector for small field proton radiosurgery.

    PubMed

    McAuley, Grant A; Teran, Anthony V; Slater, Jerry D; Slater, James M; Wroe, Andrew J

    2015-11-08

    The small fields and sharp gradients typically encountered in proton radiosurgery require high spatial resolution dosimetric measurements, especially below 1-2 cm diameters. Radiochromic film provides high resolution, but requires postprocessing and special handling. Promising alternatives are diode detectors with small sensitive volumes (SV) that are capable of high resolution and real-time dose acquisition. In this study we evaluated the PTW PR60020 proton dosimetry diode using radiation fields and beam energies relevant to radiosurgery applications. Energies of 127 and 157 MeV (9.7 to 15 cm range) and initial diameters of 8, 10, 12, and 20mm were delivered using single-stage scattering and four modulations (0, 15, 30, and 60mm) to a water tank in our treatment room. Depth dose and beam profile data were compared with PTW Markus N23343 ionization chamber, EBT2 Gafchromic film, and Monte Carlo simulations. Transverse dose profiles were measured using the diode in "edge-on" orientation or EBT2 film. Diode response was linear with respect to dose, uniform with dose rate, and showed an orientation-dependent (i.e., beam parallel to, or perpendicular to, detector axis) response of less than 1%. Diodevs. Markus depth-dose profiles, as well as Markus relative dose ratio vs. simulated dose-weighted average lineal energy plots, suggest that any LET-dependent diode response is negligible from particle entrance up to the very distal portion of the SOBP for the energies tested. Finally, while not possible with the ionization chamber due to partial volume effects, accurate diode depth-dose measurements of 8, 10, and 12 mm diameter beams were obtained compared to Monte Carlo simulations. Because of the small SV that allows measurements without partial volume effects and the capability of submillimeter resolution (in edge-on orientation) that is crucial for small fields and high-dose gradients (e.g., penumbra, distal edge), as well as negligible LET dependence over nearly the

  3. Evaluation of the dosimetric properties of a diode detector for small field proton radiosurgery

    PubMed Central

    Teran, Anthony V.; Slater, Jerry D.; Slater, James M.; Wroe, Andrew J.

    2015-01-01

    The small fields and sharp gradients typically encountered in proton radiosurgery require high spatial resolution dosimetric measurements, especially below 1–2 cm diameters. Radiochromic film provides high resolution, but requires postprocessing and special handling. Promising alternatives are diode detectors with small sensitive volumes (SV) that are capable of high resolution and real‐time dose acquisition. In this study we evaluated the PTW PR60020 proton dosimetry diode using radiation fields and beam energies relevant to radiosurgery applications. Energies of 127 and 157 MeV (9.7 to 15 cm range) and initial diameters of 8, 10, 12, and 20 mm were delivered using single‐stage scattering and four modulations (0, 15, 30, and 60 mm) to a water tank in our treatment room. Depth dose and beam profile data were compared with PTW Markus N23343 ionization chamber, EBT2 Gafchromic film, and Monte Carlo simulations. Transverse dose profiles were measured using the diode in "edge‐on" orientation or EBT2 film. Diode response was linear with respect to dose, uniform with dose rate, and showed an orientation‐dependent (i.e., beam parallel to, or perpendicular to, detector axis) response of less than 1%. Diode vs. Markus depth‐dose profiles, as well as Markus relative dose ratio vs. simulated dose‐weighted average lineal energy plots, suggest that any LET‐dependent diode response is negligible from particle entrance up to the very distal portion of the SOBP for the energies tested. Finally, while not possible with the ionization chamber due to partial volume effects, accurate diode depth‐dose measurements of 8, 10, and 12 mm diameter beams were obtained compared to Monte Carlo simulations. Because of the small SV that allows measurements without partial volume effects and the capability of submillimeter resolution (in edge‐on orientation) that is crucial for small fields and high‐dose gradients (e.g., penumbra, distal edge), as well as negligible LET

  4. Errors introduced by dose scaling for relative dosimetry

    PubMed Central

    Watanabe, Yoichi; Hayashi, Naoki

    2012-01-01

    Some dosimeters require a relationship between detector signal and delivered dose. The relationship (characteristic curve or calibration equation) usually depends on the environment under which the dosimeters are manufactured or stored. To compensate for the difference in radiation response among different batches of dosimeters, the measured dose can be scaled by normalizing the measured dose to a specific dose. Such a procedure, often called “relative dosimetry”, allows us to skip the time‐consuming production of a calibration curve for each irradiation. In this study, the magnitudes of errors due to the dose scaling procedure were evaluated by using the characteristic curves of BANG3 polymer gel dosimeter, radiographic EDR2 films, and GAFCHROMIC EBT2 films. Several sets of calibration data were obtained for each type of dosimeters, and a calibration equation of one set of data was used to estimate doses of the other dosimeters from different batches. The scaled doses were then compared with expected doses, which were obtained by using the true calibration equation specific to each batch. In general, the magnitude of errors increased with increasing deviation of the dose scaling factor from unity. Also, the errors strongly depended on the difference in the shape of the true and reference calibration curves. For example, for the BANG3 polymer gel, of which the characteristic curve can be approximated with a linear equation, the error for a batch requiring a dose scaling factor of 0.87 was larger than the errors for other batches requiring smaller magnitudes of dose scaling, or scaling factors of 0.93 or 1.02. The characteristic curves of EDR2 and EBT2 films required nonlinear equations. With those dosimeters, errors larger than 5% were commonly observed in the dose ranges of below 50% and above 150% of the normalization dose. In conclusion, the dose scaling for relative dosimetry introduces large errors in the measured doses when a large dose scaling is

  5. SU-F-T-506: Development and Commissioning of the Effective and Efficient Grid Therapy Using High Dose Rate Flattening Filter Free Beam and Multileaf Collimator

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Liu, M; Wen, N; Beyer, C

    Purpose: Treating bulky tumors with grid therapy (GT) has demonstrated high response rates. Long delivery time (∼15min), with consequent increased risk of intrafraction motion, is a major disadvantage of conventional MLC-based GT (MLC-GT). The goal of this study was to develop and commission a MLC-GT technique with similar dosimetric characteristics, but more efficient delivery. Methods: Grid plan was designed with 10X-FFF (2400MU/min) beam and MLC in a commercial treatment planning system (TPS). Grid size was 1cm by 1cm and grid-to-grid distance was 2cm. Field-in-field technique was used to flatten the dose profile at depth of 10cm. Prescription was 15Gy atmore » 1.5cm depth. Doses were verified at depths of 1.5cm, 5cm and 10cm. Point dose was measured with a plastic scintillator detector (PSD) while the planar dose was measured with calibrated Gafchromic EBT3 films in a 20cm think, 30cmx30cm solid water phantom. The measured doses were compared to the doses calculated in the treatment planning system. Percent depth dose (PDD) within the grid was also measured using EBT3 film. Five clinical cases were planned to compare beam-on time. Results: The valley-to-peak dose ratio at the 3 depths was approximately 10–15%, which is very similar to published result. The average point dose difference between the PSD measurements and TPS calculation is 2.1±0.6%. Film dosimetry revealed good agreement between the delivered and calculated dose. The average gamma passing rates at the 3 depths were 95% (3%, 1mm). The average percent difference between the measured PDD and calculated PDD was 2.1% within the depth of 20cm. The phantom plan delivery time was 3.6 min. Average beam-on time was reduced by 66.1±5.6% for the 5 clinical cases. Conclusion: An effective and efficient GT technique was developed and commissioned for the treatment of bulky tumors using FFF beam combined with MLC and automation. The Department of Radiation Oncology at Henry Ford Health System receives

  6. SU-F-BRE-03: Consideration of a Track-Interaction Model for Radiochromic Film Response

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rosen, B; DeWerd, L

    2014-06-15

    Purpose: Conventional methods for characterizing the energy response of radiochromic film (RCF) typically involve assessing changes in response when exposed to various beam qualities and use Monte Carlo to determine absorbed dose. These methods represent RCF as a dose integrator of a homogeneous energy deposition volume. Apparent film saturation, nonlinearity, and intrinsic energy dependence are unpredicted with conventional methods. Recent work has shown significant RCF intrinsic energy dependence, which limits its use in absolute dosimetry. This work introduces a track-interaction model (TIM) for RCF and assesses its ability to predict total energy response. Methods: A TIM based on Katz single-hitmore » theory was developed to accumulate energy flux along particle tracks within active crystals, represented as (1×1×20)um{sup 3} prisms about the Gafchromic™ EBT3 active volume using MCNP5 and Matlab. Energy flux contributed to film response only if near the threshold energy for polymerization in polydiacetylenes (2.5eV/monomer). Energy deposition in excess of maximum efficiency represented crystal saturation and did not contribute to film response. The TIM was applied to RCF exposed in air to various monoenergetic photon beams and Co-60. Geometric distribution of energy flux was found for each beam quality in a (1×1)mm{sup 2} RCF area. RCF response relative to Co-60 absorbed dose-to-water (S-TIM) was determined and compared to published values (S-PUB). Results: TIM successfully predicted that lower energy radiation is less effective at inducing polymerization, though the magnitude of the phenomenon was overpredicted. S-TIM was −29% and +20% for 20 and 40 keV, respectively. This agreed qualitatively with S-PUB of −27% and +16%. TIM-generated sensitometric curves contained the non-linearity and saturation apparent in RCF. Conclusion: This work indicates the possibility for TIMs to predict changes in RCF response to various energies. Future work will refine

  7. EBT Fidelity Trajectories Across Training Cohorts Using the Interagency Collaborative Team Strategy.

    PubMed

    Chaffin, Mark; Hecht, Debra; Aarons, Greg; Fettes, Danielle; Hurlburt, Michael; Ledesma, Karla

    2016-03-01

    The Interdisciplinary Collaborative Team (ICT) strategy uses front-line providers as adaptation, training and quality control agents for multi-agency EBT implementation. This study tests whether an ICT transmits fidelity to subsequent provider cohorts. SafeCare was implemented by home visitors from multiple community-based agencies contracting with child welfare. Client-reported fidelity trajectories for 5,769 visits, 957 clients and 45 providers were compared using three-level growth models. Provider cohorts trained and live-coached by the ICT attained benchmark fidelity after 12 weeks, and this was sustained. Hispanic clients reported high cultural competency, supporting a cultural adaptation crafted by the ICT.

  8. Confirmation of skin doses resulting from bolus effect of intervening alpha-cradle and carbon fiber couch in radiotherapy.

    PubMed

    Chan, Maria F; Chiu-Tsao, Sou-Tung; Li, Jingdong; Schupak, Karen; Parhar, Preeti; Burman, Chandra

    2012-12-01

    In this study, we verified the treatment planning calculations of skin doses with the incorporation of the bolus effect due to the intervening alpha-cradle (AC) and carbon fiber couch (CFC) using radiochromic EBT2 films. A polystyrene phantom (25 × 25 × 15 cm(3)) with six EBT2 films separated by polystyrene slabs, at depths of 0, 0.1, 0.2, 0.5, 1, 1.4 cm, was positioned above an AC, which was ~1 cm thick. The phantom and AC assembly were CT scanned and the CT-images were transferred to the treatment planning system (TPS) for calculations in three scenarios: (A) ignoring AC and CFC, (B) accounting for AC only, (C) accounting for both AC and CFC. A single posterior 10 × 10 cm(2) field, a pair of posterior-oblique 10 × 10 cm(2) fields, and a posterior IMRT field (6 MV photons from a Varian Trilogy linac) were planned. For each radiation field configuration, the same MU were used in all three scenarios in the TPS. Each plan for scenario C was delivered to expose a stack of EBT2 films in the phantom through AC and CFC. In addition, in vivo EBT2 film measurement on a lung cancer patient immobilized with AC undergoing IMRT was also included in this study. Point doses and planar distributions generated from the TPS for the three scenarios were compared with the data from the EBT2 film measurements. For all the field arrangements, the EBT2 film data including the in vivo measurement agreed with the doses calculated for scenario (C), within the uncertainty of the EBT2 measurements (~4%). For the single posterior field (a pair of posterior-oblique fields), the TPS generated doses were lower than the EBT2 doses by 34%, 33%, 31%, 13% (34%, 31%, 31%, 11%) for scenario A and by 27%, 25%, 22%, 8% (25%, 21%, 21%, 6%) for scenario B at the depths of 0, 0.1, 0.2, 0.5 cm, respectively. For the IMRT field, the 2D dose distributions at each depth calculated in scenario C agree with those measured data. When comparing the central axis doses for the IMRT field, we found the TPS

  9. SU-E-T-756: Tissue Inhomogeneity Corrections in Intra-Operative Radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sethi, A; Chinsky, B; Gros, S

    Purpose: Investigate the impact of tissue inhomogeneities on dose distributions produced by low-energy X-rays in intra-operative radiotherapy (IORT). Methods: A 50-kV INTRABEAM X-ray device with superficial (Flat and Surface) applicators was commissioned at our institution. For each applicator, percent depth-dose (PDD), dose-profiles (DP) and output factors (OF) were obtained. Calibrated GaFchromic (EBT3) films were used to measure dose distributions in solid water phantom at various depths (2, 5, 10, and 15 mm). All recommended precautions for film-handling, film-exposure and scanning were observed. The effects of tissue inhomogeneities on dose distributions were examined by placing air-cavities and bone and tissue equivalentmore » materials of different density (ρ), atomic number (Z), and thickness (t = 0–4mm) between applicator and film detector. All inhomogeneities were modeled as a cylindrical cavity (diameter 25 mm). Treatment times were calculated to deliver 1Gy dose at 5mm depth. Film results were verified by repeat measurements with a thin-window parallel plate ion-chamber (PTW 34013A) in a water tank. Results: For a Flat-4cm applicator, the measured dose rate at 5mm depth in solid water was 0.35 Gy/min. Introduction of a cylindrical air-cavity resulted in an increased dose past the inhomogeneity. Compared to tissue equivalent medium, dose enhancement due to 1mm, 2mm, 3mm and 4mm air cavities was 10%, 16%, 24%, and 35% respectively. X-ray attenuation by 2mm thick cortical bone resulted in a significantly large (58%) dose decrease. Conclusion: IORT dose calculations assume homogeneous tissue equivalent medium. However, soft X-rays are easily affected by non-tissue equivalent materials. The results of this study may be used to estimate and correct IORT dose delivered in the presence of tissue inhomogeneities.« less

  10. SU-E-T-439: Fundamental Verification of Respiratory-Gated Spot Scanning Proton Beam Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hamano, H; Yamakawa, T; Hayashi, N

    Purpose: The spot-scanning proton beam irradiation with respiratory gating technique provides quite well dose distribution and requires both dosimetric and geometric verification prior to clinical implementation. The purpose of this study is to evaluate the impact of gating irradiation as a fundamental verification. Methods: We evaluated field width, flatness, symmetry, and penumbra in the gated and non-gated proton beams. The respiration motion was distinguished into 3 patterns: 10, 20, and 30 mm. We compared these contents between the gated and non-gated beams. A 200 MeV proton beam from PROBEAT-III unit (Hitachi Co.Ltd) was used in this study. Respiratory gating irradiationmore » was performed by Quasar phantom (MODUS medical devices) with a combination of dedicated respiratory gating system (ANZAI Medical Corporation). For radiochromic film dosimetry, the calibration curve was created with Gafchromic EBT3 film (Ashland) on FilmQA Pro 2014 (Ashland) as film analysis software. Results: The film was calibrated at the middle of spread out Bragg peak in passive proton beam. The field width, flatness and penumbra in non-gated proton irradiation with respiratory motion were larger than those of reference beam without respiratory motion: the maximum errors of the field width, flatness and penumbra in respiratory motion of 30 mm were 1.75% and 40.3% and 39.7%, respectively. The errors of flatness and penumbra in gating beam (motion: 30 mm, gating rate: 25%) were 0.0% and 2.91%, respectively. The results of symmetry in all proton beams with gating technique were within 0.6%. Conclusion: The field width, flatness, symmetry and penumbra were improved with the gating technique in proton beam. The spot scanning proton beam with gating technique is feasible for the motioned target.« less

  11. "Edge-on" MOSkin detector for stereotactic beam measurement and verification.

    PubMed

    Jong, Wei Loong; Ung, Ngie Min; Vannyat, Ath; Jamalludin, Zulaikha; Rosenfeld, Anatoly; Wong, Jeannie Hsiu Ding

    2017-01-01

    Dosimetry in small radiation field is challenging and complicated because of dose volume averaging and beam perturbations in a detector. We evaluated the suitability of the "Edge-on" MOSkin (MOSFET) detector in small radiation field measurement. We also tested the feasibility for dosimetric verification in stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT). "Edge-on" MOSkin detector was calibrated and the reproducibility and linearity were determined. Lateral dose profiles and output factors were measured using the "Edge-on" MOSkin detector, ionization chamber, SRS diode and EBT2 film. Dosimetric verification was carried out on two SRS and five SRT plans. In dose profile measurements, the "Edge-on" MOSkin measurements concurred with EBT2 film measurements. It showed full width at half maximum of the dose profile with average difference of 0.11mm and penumbral width with difference of ±0.2mm for all SRS cones as compared to EBT2 film measurement. For output factor measurements, a 1.1% difference was observed between the "Edge-on" MOSkin detector and EBT2 film for 4mm SRS cone. The "Edge-on" MOSkin detector provided reproducible measurements for dose verification in real-time. The measured doses concurred with the calculated dose for SRS (within 1%) and SRT (within 3%). A set of output correction factors for the "Edge-on" MOSkin detector for small radiation fields were derived from EBT2 film measurement and presented. This study showed that the "Edge-on" MOSkin detector is a suitable tool for dose verification in small radiation field. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  12. EBT Fidelity Trajectories Across Training Cohorts Using the Interagency Collaborative Team Strategy

    PubMed Central

    Hecht, Debra; Aarons, Greg; Fettes, Danielle; Hurlburt, Michael; Ledesma, Karla

    2015-01-01

    The Interdisciplinary Collaborative Team (ICT) strategy uses front-line providers as adaptation, training and quality control agents for multi-agency EBT implementation. This study tests whether an ICT transmits fidelity to subsequent provider cohorts. SafeCare was implemented by home visitors from multiple community-based agencies contracting with child welfare. Client-reported fidelity trajectories for 5,769 visits, 957 clients and 45 providers were compared using three-level growth models. Provider cohorts trained and live-coached by the ICT attained benchmark fidelity after 12 weeks, and this was sustained. Hispanic clients reported high cultural competency, supporting a cultural adaptation crafted by the ICT. PMID:25586878

  13. A respiratory compensating system: design and performance evaluation

    PubMed Central

    Huang, Ding‐Yang; Tien, Der‐Chi; Wu, Ren‐Hong; Hsu, Chung‐Hsien

    2014-01-01

    This study proposes a respiratory compensating system which is mounted on the top of the treatment couch for reverse motion, opposite from the direction of the targets (diaphragm and hemostatic clip), in order to offset organ displacement generated by respiratory motion. Traditionally, in the treatment of cancer patients, doctors must increase the field size for radiation therapy of tumors because organs move with respiratory motion, which causes radiation‐induced inflammation on the normal tissues (organ at risk (OAR)) while killing cancer cells, and thereby reducing the patient's quality of life. This study uses a strain gauge as a respiratory signal capture device to obtain abdomen respiratory signals, a proposed respiratory simulation system (RSS) and respiratory compensating system to experiment how to offset the organ displacement caused by respiratory movement and compensation effect. This study verifies the effect of the respiratory compensating system in offsetting the target displacement using two methods. The first method uses linac (medical linear accelerator) to irradiate a 300 cGy dose on the EBT film (GAFCHROMIC EBT film). The second method uses a strain gauge to capture the patients' respiratory signals, while using fluoroscopy to observe in vivo targets, such as a diaphragm, to enable the respiratory compensating system to offset the displacements of targets in superior‐inferior (SI) direction. Testing results show that the RSS position error is approximately 0.45 ~ 1.42 mm, while the respiratory compensating system position error is approximately 0.48 ~ 1.42 mm. From the EBT film profiles based on different input to the RSS, the results suggest that when the input respiratory signals of RSS are sine wave signals, the average dose (%) in the target area is improved by 1.4% ~ 24.4%, and improved in the 95% isodose area by 15.3% ~ 76.9% after compensation. If the respiratory signals input into the RSS respiratory signals are actual human

  14. Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry

    PubMed Central

    Lewis, David; Chan, Maria F.

    2015-01-01

    Purpose: A known factor affecting the accuracy of radiochromic film dosimetry is the lateral response artifact (LRA) induced by nonuniform response of a flatbed scanner in the direction perpendicular to the scan direction. This work reports a practical solution to eliminate such artifacts for all forms of dose QA. Methods: EBT3 films from a single production lot (02181401) cut into rectangular 4 × 5 cm2 pieces, with the long dimension parallel to the long dimension of the original 20.3 × 25.4 cm2 sheets, were exposed at a depth of 5 cm on a Varian Trilogy at the center of a 20 × 20 cm2 open field at seven doses between 50 and 1600 cGy using 6 MV photons. These films together with an unexposed film from the same production lot were lined one next to the other on an Epson 10000XL or 11000XL scanner in portrait orientation with their long dimension parallel to the scan direction. Scanned images were then obtained with the line of films positioned at seven discrete lateral locations perpendicular to the scan direction. The process was repeated in landscape orientation and on three other Epson scanners. Data were also collected for three additional production lots of EBT3 film (11051302, 03031401, and 03171403). From measurements at the various lateral positions, the scanner response was determined as a function of the lateral position of the scanned film. For a given color channel X, the response at any lateral position L is related to the response at the center, C, of the scanner by Response(C, D, X) = AL,X + BL,X⋅Response(L, D, X), where D is dose and the coefficients AL,X and BL,X are determined from the film measurements at the center of the scanner and six other discrete lateral positions. The values at intermediate lateral positions were obtained by linear interpolation. The coefficients were determined for the red, green, and blue color channels, preserving the ability to apply triple-channel dosimetry once corrections were applied to compensate for the

  15. EBT Payment for Online Grocery Orders: a Mixed-Methods Study to Understand Its Uptake among SNAP Recipients and the Barriers to and Motivators for Its Use.

    PubMed

    Martinez, Olivia; Tagliaferro, Barbara; Rodriguez, Noemi; Athens, Jessica; Abrams, Courtney; Elbel, Brian

    2018-04-01

    To examine Supplemental Nutrition Assistance Program (SNAP) recipients' use of the first online supermarket accepting Electronic Benefit Transfer (EBT) payment. In this mixed-methods study, the authors collected EBT purchase data from an online grocer and attempted a randomized controlled trial in the South Bronx, New York City, followed by focus groups with SNAP beneficiaries aged ≥18 years. Participants were randomized to shop at their usual grocery store or an online supermarket for 3 months. Focus groups explored barriers and motivators to online EBT redemption. Few participants made online purchases, even when incentivized in the randomized controlled trial. Qualitative findings highlighted a lack of perceived control over the online food selection process as a key barrier to purchasing food online. Motivators included fast, free shipping and discounts. Electronic Benefit Transfer for online grocery purchases has the potential to increase food access among SNAP beneficiaries, but challenges exist to this new food buying option. Understanding online food shopping barriers and motivators is critical to the success of policies targeting the online expansion of SNAP benefits. Copyright © 2017 Society for Nutrition Education and Behavior. Published by Elsevier Inc. All rights reserved.

  16. Experimental and Monte Carlo measurements of dose perturbation around a non-radioactive brachytherapy seed in external beam radiotherapy

    NASA Astrophysics Data System (ADS)

    Steinman, James P.

    I-125 seeds used in permanent prostate brachytherapy are composed of high-Z metals and may number from 40 to over 100 in a typical implant. If any supplemental external beam treatment is administered afterward (as for salvaging failed brachytherapy treatment), it is possible that the seeds may cause substantial dose perturbation which will depend on numerous factors (photon energy, depth, field size, number of seeds, etc.) and this effect needs to be thoroughly investigated. Film measurements were primarily done using Kodak XV2 layered above and below a non-radioactive I-125 seed placed in a groove on a Lucite plate with 5 cm buildup and 10 cm backscatter added at 95 cm SSD. The phantom was irradiated with and without seed with 6 MV photons for a 1 x 1 cm2 field size. Monte Carlo simulations were carried out using DOSXYZnrc using the same parameters and compared with Gafchromic EBT2 film. Other comparisons looked at changing energy, depth, and field size in both with and without seeds configuration. This study was further extended to include metals of various Z of the seed's dimensions and also looked into effect of 3 seeds spaced 0.5 cm vertically. Another measurement was done using two opposing fields using single as well as 3 seed configuration to see whether the dose enhancement and attenuation cancel out in multi-field treatments which is the norm clinically in a prostate treatment. For a single I-125 seed, on XV film a localized dose enhancement of 6.3% upstream and -10.9% downstream was noticed. With three seeds, this effect did not change. With two opposing fields, a cold spot around the seed of ~3% was noticed from film measurements. Increasing energy and field size decreased the effect while increase in Z of material greatly increased the effect. Increasing depth appeared to have no effect. DOSXYZnrc and EBT2 film verified maximum dose enhancement of +15% upstream and -20% downstream of the I-125 seed surface. In general, the range of the effect was

  17. Dosimetry of a Small-Animal Irradiation Model using a 6 MV Linear Accelerator

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fitch, F. Moran; Martinez-Davalos, A.; Garcia-Garduno, O. A.

    2010-12-07

    A custom made rat-like phantom was used to measure dose distributions using a 6 MV linear accelerator. The phantom has air cavities that simulate the lungs and cylindrical inserts that simulate the backbone. The calculated dose distributions were obtained with the BrainScan v.5.31 TPS software. For the irradiation two cases were considered: (a) near the region where the phantom has two air cavities that simulate the lungs, and (b) with an entirely uniform phantom. The treatment plan consisted of two circular cone arcs that imparted a 500 cGy dose to a simulated lesion in the backbone. We measured dose distributionsmore » using EBT2 GafChromic film and an Epson Perfection V750 scanner working in transmission mode. Vertical and horizontal profiles, isodose curves from 50 to 450 cGy, dose and distance to agreement (DTA) histograms and Gamma index were obtained to compare the dose distributions using DoseLab v4.11. As a result, these calculations show very good agreement between calculated and measured dose distribution in both cases. With a 2% 2 mm criteria 100% of the points pass the Gamma test for the uniform case, while 98.9% of the points do it for the lungs case.« less

  18. Evaluation of effective energy for QA and QC: measurement of half-value layer using radiochromic film density.

    PubMed

    Gotanda, T; Katsuda, T; Gotanda, R; Tabuchi, A; Yamamoto, K; Kuwano, T; Yatake, H; Takeda, Y

    2009-03-01

    The effective energy of diagnostic X-rays is important for quality assurance (QA) and quality control (QC). However, the half-value layer (HVL), which is necessary to evaluate the effective energy, is not ubiquitously monitored because ionization-chamber dosimetry is time-consuming and complicated. To verify the applicability of GAFCHROMIC XR type R (GAF-R) film for HVL measurement as an alternative to monitoring with an ionization chamber, a single-strip method for measuring the HVL has been evaluated. Calibration curves of absorbed dose versus film density were generated using this single-strip method with GAF-R film, and the coefficient of determination (r2) of the straight-line approximation was evaluated. The HVLs (effective energies) estimated using the GAF-R film and an ionization chamber were compared. The coefficient of determination (r2) of the straight-line approximation obtained with the GAF-R film was more than 0.99. The effective energies (HVLs) evaluated using the GAF-R film and the ionization chamber were 43.25 keV (5.10 mm) and 39.86 keV (4.45 mm), respectively. The difference in the effective energies determined by the two methods was thus 8.5%. These results suggest that GAF-R might be used to evaluate the effective energy from the film-density growth without the need for ionization-chamber measurements.

  19. Development of radiochromic film for spatially quantitative dosimetric analysis of indirect ionizing radiation fields

    NASA Astrophysics Data System (ADS)

    Brady, Samuel Loren

    Two types of radiochromic films (RCF) were characterized for this work: EBT and XRQA film. Both films were investigated for: radiation interaction with film structure; light interaction with film structure for optimal film readout (densitometry) sensitivity; range of absorbed dose measurements; dependence of film dose measurement response as a function of changing radiation energy; fractionation and dose rate effects on film measurement response; film response sensitivity to ambient factors; and stability of measured film response with time. EBT film was shown to have the following properties: near water equivalent atomic weight (Zeff); dynamic dose range of 10 -1-102 Gy; 3% change in optical density (OD) response for a single exposure level when exposed to radiation energies from (75-18,000) kV; and best digitized using transmission densitometry. XRQA film was shown to have: a Zeff of ˜25; a 12 fold increase in sensitivity at lower photon energies for a dynamic dose range of 10-3-100 Gy, a difference of 25% in OD response when comparing 120 kV to 320 kV, and best digitized using reflective densitometry. Both XRQA and EBT films were shown to have: a temporal stability (DeltaOD) of ˜1% for t > 24 hr post film exposure for up to ˜20 days; a change in dose response of ˜0.03 mGy hr-1 when exposed to fluorescent room lighting at standard room temperature and humidity levels; a negligible dose rate and fractionation effect when operated within the optimal dose ranges; and a light wavelength dependence with dose for film readout. The flat bed scanner was chosen as the primary film digitizer due to its availability, cost, OD range, functionality (transmission and reflection scanning), and digitization speed. As a cost verses functionality comparison, the intrinsic and operational limitations were determined for two flat bed scanners. The EPSON V700 and 10000XL exhibited equal spatial and OD accuracy. The combined precision of both the scanner light sources and CCD

  20. Radiochromic film calibration for the RQT9 quality beam

    NASA Astrophysics Data System (ADS)

    Costa, K. C.; Gomez, A. M. L.; Alonso, T. C.; Mourao, A. P.

    2017-11-01

    When ionizing radiation interacts with matter it generates energy deposition. Radiation dosimetry is important for medical applications of ionizing radiation due to the increasing demand for diagnostic radiology and radiotherapy. Different dosimetry methods are used and each one has its advantages and disadvantages. The film is a dose measurement method that records the energy deposition by the darkening of its emulsion. Radiochromic films have a little visible light sensitivity and respond better to ionizing radiation exposure. The aim of this study is to obtain the resulting calibration curve by the irradiation of radiochromic film strips, making it possible to relate the darkening of the film with the absorbed dose, in order to measure doses in experiments with X-ray beam of 120 kV, in computed tomography (CT). Film strips of GAFCHROMIC XR-QA2 were exposed according to RQT9 reference radiation, which defines an X-ray beam generated from a voltage of 120 kV. Strips were irradiated in "Laboratório de Calibração de Dosímetros do Centro de Desenvolvimento da Tecnologia Nuclear" (LCD / CDTN) at a dose range of 5-30 mGy, corresponding to the range values commonly used in CT scans. Digital images of the irradiated films were analyzed by using the ImageJ software. The darkening responses on film strips according to the doses were observed and they allowed obtaining the corresponding numeric values to the darkening for each specific dose value. From the numerical values of darkening, a calibration curve was obtained, which correlates the darkening of the film strip with dose values in mGy. The calibration curve equation is a simplified method for obtaining absorbed dose values using digital images of radiochromic films irradiated. With the calibration curve, radiochromic films may be applied on dosimetry in experiments on CT scans using X-ray beam of 120 kV, in order to improve CT acquisition image processes.

  1. Calibration of GafChromic XR-RV3 radiochromic film for skin dose measurement using standardized x-ray spectra and a commercial flatbed scanner

    PubMed Central

    McCabe, Bradley P.; Speidel, Michael A.; Pike, Tina L.; Van Lysel, Michael S.

    2011-01-01

    Purpose: In this study, newly formulated XR-RV3 GafChromic® film was calibrated with National Institute of Standards and Technology (NIST) traceability for measurement of patient skin dose during fluoroscopically guided interventional procedures. Methods: The film was calibrated free-in-air to air kerma levels between 15 and 1100 cGy using four moderately filtered x-ray beam qualities (60, 80, 100, and 120 kVp). The calibration films were scanned with a commercial flatbed document scanner. Film reflective density-to-air kerma calibration curves were constructed for each beam quality, with both the orange and white sides facing the x-ray source. A method to correct for nonuniformity in scanner response (up to 25% depending on position) was developed to enable dose measurement with large films. The response of XR-RV3 film under patient backscattering conditions was examined using on-phantom film exposures and Monte Carlo simulations. Results: The response of XR-RV3 film to a given air kerma depended on kVp and film orientation. For a 200 cGy air kerma exposure with the orange side of the film facing the source, the film response increased by 20% from 60 to 120 kVp. At 500 cGy, the increase was 12%. When 500 cGy exposures were performed with the white side facing the x-ray source, the film response increased by 4.0% (60 kVp) to 9.9% (120 kVp) compared to the orange-facing orientation. On-phantom film measurements and Monte Carlo simulations show that using a NIST-traceable free-in-air calibration curve to determine air kerma in the presence of backscatter results in an error from 2% up to 8% depending on beam quality. The combined uncertainty in the air kerma measurement from the calibration curves and scanner nonuniformity correction was ±7.1% (95% C.I.). The film showed notable stability. Calibrations of film and scanner separated by 1 yr differed by 1.0%. Conclusions: XR-RV3 radiochromic film response to a given air kerma shows dependence on beam quality and film

  2. Spectrophotometric determination of β-adrenergic antagonists drugs via ion-pair complex formation using MO and EBT

    NASA Astrophysics Data System (ADS)

    El-Didamony, A. M.; Shehata, A. M.

    2014-09-01

    Two simple, rapid and sensitive spectrophotometric methods have been proposed for the assay of bisoprolol fumarate (BSF), propranolol hydrochloride (PRH), and timolol maleate (TIM) either in bulk or in pharmaceutical formulations. The methods are based on the reaction of the selected drugs with methyl orange (MO) and eriochrome black T in acidic buffers, after extracting in dichloromethane and measured quantitatively with maximum absorption at 428 and 518 nm for MO and EBT, respectively. The analytical parameters and their effects on the reported systems are investigated. The extracts are intensely colored and very stable at room temperature. The calibration graphs were linear over the concentration range of 0.8-6.4, 0.4-3.6, 0.8-5.6 μg/mL for BSF, PRH, and TIM, respectively, with MO and 0.8-6.4, 0.4-3.2, and 0.8-8.0 μg/mL for BSF, PRH, and TIM, respectively, with EBT. The stoichiometry of the complexes was found to be 1 : 1 in all cases. The proposed methods were successfully extended to pharmaceutical preparations. Excipients used as additive in commercial formulations did not interfere in the analysis. The proposed methods can be recommended for quality control and routine analysis where time, cost effectiveness and high specificity of analytical technique are of great importance.

  3. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Saleh, H; Ferjani, S; Masssey, V

    Purpose: Perform dosimetric comparison between planned and delivered dose in the junction area, measure daily dose variation in the arc junction area for pediatric patients treated for medulloblastoma using Craniospinal axis irradiation(CSI) Material and methods Dose comparison in the junction area, daily dose variation in the arc junction area for a Rando Phantom and 5 pediatric patients treated using CSI technique were analyzed. Plans were created using the Eclipse treatment planning system. Two arcs for cranium and 1 arc for spine region were used. Planar dose matrix was created by projecting phantom and patient plan into the ArcCheck phantom. EBT3more » film was placed in the middle of ArcCheck plug to measure dose distribution in the junction areaDuring patient treatment, strip of EBT3 film was placed daily at each junction area for verification. EBT3 films were scanned using a flatbed scanner, Epson Expression 10000 XL. Film QA pro software was used to analyze film. Scanning and analysis was performed according to vendor recommendations and AAPM TG-55 report. Films were scanned and analyzed daily after each treatment and at the end of treatment course. Planar dose distributions from films were compared with planar dose distribution from treatment planning system. Results: Comparison of planned vs. measured dose distributions for patients have passing rates of 90%–100% with 3% and 3 mm gamma analysis. In some of the treatment fractions, daily setup film showed variation in dose distribution in the junction area. Conclusion: It is critical to measure dose distribution in the arc junction area and use additional quality assurance measures to verify daily setup for CSI patient where one or more junctions are present. EBT3 film prove to be a good tool to achieve this task considering flexibility associated with the film such as symmetry, self-developing and ease of use.« less

  4. SU-G-TeP2-07: Dosimetric Characterization of a New HDR Multi-Channel Esophageal Applicator for Brachytherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zhao, A; Gao, S; Greskovich, J

    2016-06-15

    Purpose: To characterize the dose distribution of a new multi-channel esophageal applicator for brachytherapy HDR treatment, and particularly the effect of the presence of air or water in the applicator’s expansion balloon. Methods: A new multi-channel (6) inflatable applicator for esophageal HDR has been developed in house and tested in a simple water phantom. CT image sets were obtained under several balloon expansions (80ml of air, 50 cc of water), and channel loadings and used with the Oncentra (Elekta) planning system based on TG43 formalism. 400 cGy was prescribed to a plane 1cm away from the applicator. Planar dose distributionsmore » were measured for that plane and one next to the applicator using Gafchromic EBT3 film and scanned by a Vidar VXR-12 film digitizer. Film and TPS generated dose distributions of film were sent to OmniPro I’mRT (iba DOSIMETRY) for analysis. 2D dose profiles in both X and Y directions were compared and gamma analysis performed. Results: Film dose measurement of the air-inflated applicator is lower than the TPS calculated dose by as much as 60%. Only 80.8% of the pixels passed the gamma criteria (3%/3mm). For the water-inflated applicator, the measured film dose is fairly close to the TPS calculated dose (typically within <3%). 99.84% of the pixels passed the gamma criteria (3%/3mm). Conclusion: TG43 based calculations worked well when water was used in the expansion balloon. However, when air is present in that balloon, the neglect of heterogeneity corrections in the TG43 calculation results in large differences between calculated and measured doses. This could result in severe underdosing when used in a patient. This study illustrates the need for a TPS with an advanced algorithm which can account for heterogeneity. Supported by Innovations Department, Cleveland Clinic.« less

  5. Experimental evaluation of a GPU-based Monte Carlo dose calculation algorithm in the Monaco treatment planning system.

    PubMed

    Paudel, Moti R; Kim, Anthony; Sarfehnia, Arman; Ahmad, Sayed B; Beachey, David J; Sahgal, Arjun; Keller, Brian M

    2016-11-08

    A new GPU-based Monte Carlo dose calculation algorithm (GPUMCD), devel-oped by the vendor Elekta for the Monaco treatment planning system (TPS), is capable of modeling dose for both a standard linear accelerator and an Elekta MRI linear accelerator. We have experimentally evaluated this algorithm for a standard Elekta Agility linear accelerator. A beam model was developed in the Monaco TPS (research version 5.09.06) using the commissioned beam data for a 6 MV Agility linac. A heterogeneous phantom representing several scenarios - tumor-in-lung, lung, and bone-in-tissue - was designed and built. Dose calculations in Monaco were done using both the current clinical Monte Carlo algorithm, XVMC, and the new GPUMCD algorithm. Dose calculations in a Pinnacle TPS were also produced using the collapsed cone convolution (CCC) algorithm with heterogeneity correc-tion. Calculations were compared with the measured doses using an ionization chamber (A1SL) and Gafchromic EBT3 films for 2 × 2 cm2, 5 × 5 cm2, and 10 × 10 cm2 field sizes. The percentage depth doses (PDDs) calculated by XVMC and GPUMCD in a homogeneous solid water phantom were within 2%/2 mm of film measurements and within 1% of ion chamber measurements. For the tumor-in-lung phantom, the calculated doses were within 2.5%/2.5 mm of film measurements for GPUMCD. For the lung phantom, doses calculated by all of the algorithms were within 3%/3 mm of film measurements, except for the 2 × 2 cm2 field size where the CCC algorithm underestimated the depth dose by ~ 5% in a larger extent of the lung region. For the bone phantom, all of the algorithms were equivalent and calculated dose to within 2%/2 mm of film measurements, except at the interfaces. Both GPUMCD and XVMC showed interface effects, which were more pronounced for GPUMCD and were comparable to film measurements, whereas the CCC algorithm showed these effects poorly. © 2016 The Authors.

  6. Marvin: an anatomical phantom for dosimetric evaluation of complex radiotherapy of the head and neck.

    PubMed

    Aitkenhead, A H; Rowbottom, C G; Mackay, R I

    2013-10-07

    We report on the design of Marvin, a Model Anatomy for Radiotherapy Verification and audit In the head and Neck and present results demonstrating its use in the development of the Elekta volumetric modulated arc therapy (VMAT) technique at the Christie, and in the audit of TomoTherapy and Varian RapidArc at other institutions. The geometry of Marvin was generated from CT datasets of eight male and female patients lying in the treatment position, with removable inhomogeneities modelling the sinuses and mandible. A modular system allows the phantom to be used with a range of detectors, with the locations of the modules being based on an analysis of a range of typical treatment plans (27 in total) which were mapped onto the phantom geometry. Results demonstrate the use of Gafchromic EBT2/EBT3 film for measurement of relative dose in a plane through the target and organs-at-risk, and the use of a small-volume ionization chamber for measurement of absolute dose in the target and spinal cord. Measurements made during the development of the head and neck VMAT protocol at the Christie quantified the improvement in plan delivery resulting from the installation of the Elekta Integrity upgrade (which permits an effectively continuously variable dose rate), with plans delivered before and after the upgrade having 88.5 ± 9.4% and 98.0 ± 2.2% respectively of points passing a gamma analysis (at 4%, 4 mm, global). Audits of TomoTherapy and Varian RapidArc neck techniques at other institutions showed a similar quality of plan delivery as for post-Integrity Elekta VMAT: film measurements for both techniques had >99% of points passing a gamma analysis at the clinical criteria of 4%, 4 mm, global, and >95% of points passing at tighter criteria of 3%, 3 mm, global; and absolute dose measurements in the PTV and spinal cord were within 1.5% and 3.5% of the planned doses respectively for both techniques. The results demonstrate that Marvin is an efficient and effective means of

  7. SU-C-213-02: Characterizing 3D Printing in the Fabrication of Variable Density Phantoms

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Madamesila, J; McGeachy, P; Villarreal-Barajas, J

    Purpose: In this work, we present characterization, process flow, quality control and application of 3D fabricated low density phantoms for radiotherapy quality assurance. Methods: A Rostock delta 3D printer using polystyrene filament of diameter 1.75 mm was used to print geometric volumes of 2×2×1 cm{sup 3} of varying densities. The variable densities of 0.1 to 0.75 g/cm {sup 3} were created by modulating the infill. A computed tomography (CT) scan was performed to establish an infill-density calibration curve as well as characterize the quality of the print such as uniformity and the infill pattern. The time required to print thesemore » volumes was also recorded. Using the calibration, two low density cones (0.19, 0.52 g/cm{sup 3}) were printed and benchmarked against commercially available phantoms. The dosimetric validation of the low density scaling of Anisotropic Analytical Algorithm (AAA) was performed by using a 0.5 g/cm{sup 3} slab of 10×10×2.4 cm{sup 3} with EBT3 GafChromic film. The gamma analysis at 3%/3mm criteria were compared for the measured and computed dose planes. Results: Analysis of the volume of air pockets in the infill resulted in a reasonable uniformity for densities 0.4 to 0.75 g/cm{sup 3}. Printed phantoms with densities below 0.4 g/cm{sup 3} exhibited a higher ratio of air to polystyrene resulting in large non-uniformity. Compared to the commercial inserts, good agreement was observed only for the printed 0.52 g/cm{sup 3} cone. Dosimetric comparison for a printed low density volume placed in-between layers of solid water resulted in >95% gamma agreement between AAA calculated dose planes and measured EBT3 films for a 6MV 5×5 cm{sup 2} clinical beam. The comparison showed disagreement in the penumbra region. Conclusion: In conclusion, 3D printing technology opens the door to desktop fabrication of variable density phantoms at economical prices in an efficient manner for the quality assurance needs of a small clinic.« less

  8. Marvin: an anatomical phantom for dosimetric evaluation of complex radiotherapy of the head and neck

    NASA Astrophysics Data System (ADS)

    Aitkenhead, A. H.; Rowbottom, C. G.; Mackay, R. I.

    2013-10-01

    We report on the design of Marvin, a Model Anatomy for Radiotherapy Verification and audit In the head and Neck and present results demonstrating its use in the development of the Elekta volumetric modulated arc therapy (VMAT) technique at the Christie, and in the audit of TomoTherapy and Varian RapidArc at other institutions. The geometry of Marvin was generated from CT datasets of eight male and female patients lying in the treatment position, with removable inhomogeneities modelling the sinuses and mandible. A modular system allows the phantom to be used with a range of detectors, with the locations of the modules being based on an analysis of a range of typical treatment plans (27 in total) which were mapped onto the phantom geometry. Results demonstrate the use of Gafchromic EBT2/EBT3 film for measurement of relative dose in a plane through the target and organs-at-risk, and the use of a small-volume ionization chamber for measurement of absolute dose in the target and spinal cord. Measurements made during the development of the head and neck VMAT protocol at the Christie quantified the improvement in plan delivery resulting from the installation of the Elekta Integrity upgrade (which permits an effectively continuously variable dose rate), with plans delivered before and after the upgrade having 88.5 ± 9.4% and 98.0 ± 2.2% respectively of points passing a gamma analysis (at 4%, 4 mm, global). Audits of TomoTherapy and Varian RapidArc neck techniques at other institutions showed a similar quality of plan delivery as for post-Integrity Elekta VMAT: film measurements for both techniques had >99% of points passing a gamma analysis at the clinical criteria of 4%, 4 mm, global, and >95% of points passing at tighter criteria of 3%, 3 mm, global; and absolute dose measurements in the PTV and spinal cord were within 1.5% and 3.5% of the planned doses respectively for both techniques. The results demonstrate that Marvin is an efficient and effective means of

  9. Poster — Thur Eve — 20: CTDI Measurements using a Radiochromic Film-based clinical protocol

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Quintero, C.; Bekerat, H.; DeBlois, F.

    2014-08-15

    The purpose of the study was evaluating accuracy and reproducibility of a radiochromic film-based protocol to measure computer tomography dose index (CTDI) as a part of annual QA on CT scanners and kV-CBCT systems attached to linear accelerators. Energy dependence of Gafchromic XR-QA2 ® film model was tested over imaging beam qualities (50 – 140 kVp). Film pieces were irradiated in air to known values of air-kerma (up to 10 cGy). Calibration curves for each beam quality were created (Film reflectance change Vs. Air-kerma in air). Film responses for same air-kerma values were compared. Film strips were placed into holesmore » of a CTDI phantom and irradiated for several clinical scanning protocols. Film reflectance change was converted into dose to water and used to calculate CTDIvol values. Measured and tabulated CTDIvol values were compared. Average variations of ±5.2% in the mean film reflectance change were observed in the energy range of 80 to 140 keV, and 11.1% between 50 and 140 keV. Measured CTDI values were in average 10% lower than tabulated CTDI values for CT-simulators, and 44% higher for CBCT systems. Results presented a mean variation for the same machine and protocol of 2.6%. Variation of film response is within ±5% resulting in ±15% systematic error in dose estimation if a single calibration curve is used. Relatively large discrepancy between measured and tabulated CTDI values strongly support the trend towards replacing CTDI value with equilibrium dose measurement in the center of cylindrical phantom, as suggested by TG- 111.« less

  10. Dosimetric evaluation of a novel high dose rate (HDR) intraluminal / interstitial brachytherapy applicator for gastrointestinal and bladder cancers

    PubMed Central

    Aghamiri, Seyyed Mahmoud Reza; Najarian, Siamak; Jaberi, Ramin

    2010-01-01

    High dose rate (HDR) brachytherapy is one of the accepted treatment modalities in gastro‐intestinal tract and bladder carcinomas. Considering the shortcoming of contact brachytherapy routinely used in gastrointestinal tract in treatment of big tumors or invasive method of bladder treatment, an intraluminal applicator with the capability of insertion into the tumor depth seems to be useful. This study presents some dosimetric evaluations to introduce this applicator to the clinical use. The radiation attenuation characteristics of the applicator were evaluated by means of two dosimetric methods including well‐type chamber and radiochromic film. The proposed 110 cm long applicator has a flexible structure made of stainless steel for easy passage through lumens and a needle tip to drill into big tumors. The 2 mm diameter of the applicator is thick enough for source transition, while easy passage through any narrow lumen such as endoscope or cystoscope working channel is ensured. Well‐chamber results showed an acceptably low attenuation of this steel springy applicator. Performing absolute dosimetry resulted in a correlation coefficient of R=0.9916(p‐value≈10−7) between standard interstitial applicator and the one proposed in this article. This study not only introduces a novel applicator with acceptable attenuation but also proves the response independency of the GAFCHROMIC EBT films to energy. By applying the dose response of the applicator in the treatment planning software, it can be used as a new intraluminal / interstitial applicator. PACS number: 87.53.Bn, 87.53.Jw, 29.40.Cs

  11. TU-F-201-00: Radiochromic Film Dosimetry Update

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    NONE

    Since the introduction of radiochromic films (RCF) for radiation dosimetry, the scope of RCF dosimetry has expanded steadily to include many medical applications, such as radiation therapy and diagnostic radiology. The AAPM Task Group (TG) 55 published a report on the recommendations for RCF dosimetry in 1998. As the technology is advancing rapidly, and its routine clinical use is expanding, TG 235 has been formed to provide an update to TG-55 on radiochromic film dosimetry. RCF dosimetry applications in clinical radiotherapy have become even more widespread, expanding from primarily brachytherapy and radiosurgery applications, and gravitating towards (but not limited to)more » external beam therapy (photon, electron and protons), such as quality assurance for IMRT, VMAT, Tomotherapy, SRS/SRT, and SBRT. In addition, RCF applications now extend to measurements of radiation dose in particle beams and patients undergoing medical exams, especially fluoroscopically guided interventional procedures and CT. The densitometers/scanners used for RCF dosimetry have also evolved from the He-Ne laser scanner to CCD-based scanners, including roller-based scanner, light box-based digital camera, and flatbed color scanner. More recently, multichannel RCF dosimetry introduced a new paradigm for external beam dose QA for its high accuracy and efficiency. This course covers in detail the recent advancements in RCF dosimetry. Learning Objectives: Introduce the paradigm shift on multichannel film dosimetry Outline the procedures to achieve accurate dosimetry with a RCF dosimetry system Provide comprehensive guidelines on RCF dosimetry for various clinical applications One of the speakers has a research agreement from Ashland Inc., the manufacturer of Gafchromic film.« less

  12. 49 CFR 40.243 - What is the procedure for an alcohol screening test using an EBT or non-evidential breath ASD?

    Code of Federal Regulations, 2011 CFR

    2011-10-01

    ... 49 Transportation 1 2011-10-01 2011-10-01 false What is the procedure for an alcohol screening test using an EBT or non-evidential breath ASD? 40.243 Section 40.243 Transportation Office of the...-evidential breath ASD? As the BAT or STT, you must take the following steps: (a) Select, or allow the...

  13. 49 CFR 40.243 - What is the procedure for an alcohol screening test using an EBT or non-evidential breath ASD?

    Code of Federal Regulations, 2010 CFR

    2010-10-01

    ... 49 Transportation 1 2010-10-01 2010-10-01 false What is the procedure for an alcohol screening test using an EBT or non-evidential breath ASD? 40.243 Section 40.243 Transportation Office of the...-evidential breath ASD? As the BAT or STT, you must take the following steps: (a) Select, or allow the...

  14. SU-F-T-76: Total Skin Electron Therapy: An-End-To-End Examination of the Absolute Dosimetry with a Rando Phantom

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Cui, G; Ha, J; Zhou, S

    Purpose: To examine and validate the absolute dose for total skin electron therapy (TSET) through an end-to-end test with a Rando phantom using optically stimulated luminescent dosimeters (OSLDs) and EBT3 radiochromic films. Methods: A Varian Trilogy linear accelerator equipped with the special procedure 6 MeV HDTSe- was used to perform TSET irradiations using a modified Stanford 6-dual-field technique. The absolute dose was calibrated using a Markus ion chamber at a reference depth of 1.3cm at 100 cm SSD with a field size of 36 × 36 cm at the isocenter in solid water slabs. The absolute dose was cross validatedmore » by a farmer ion chamber. Then the dose rate in the unit of cGy/Mu was calibrated using the Markus chamber at the treatment position. OSLDs were used to independently verify the dose using the calibrated dose rate. Finally, a patient treatment plan (200 cGy/cycle) was delivered in the QA mode to a Rando phantom, which had 16 pairs of OSLDs and EBT3 films taped onto its surface at different anatomical positions. The doses recorded were read out to validate the absolute dosimetry for TSET. Results: The OSLD measurements were within 7% agreement with the planned dose except the shoulder areas, where the doses recorded were 23% lower on average than those of the planned. The EBT3 film measurements were within 10% agreement with the planned dose except the shoulder and the scalp vertex areas, where the respective doses recorded were 18% and 14% lower on average than those of the planned. The OSLDs gave more consistent dose measurements than those of the EBT3 films. Conclusion: The absolute dosimetry for TSET was validated by an end-to-end test with a Rando phantom using the OSLDs and EBT3 films. The beam calibration and monitor unit calculations were confirmed.« less

  15. Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry.

    PubMed

    Lewis, David; Chan, Maria F

    2015-01-01

    A known factor affecting the accuracy of radiochromic film dosimetry is the lateral response artifact (LRA) induced by nonuniform response of a flatbed scanner in the direction perpendicular to the scan direction. This work reports a practical solution to eliminate such artifacts for all forms of dose QA. EBT3 films from a single production lot (02181401) cut into rectangular 4 × 5 cm(2) pieces, with the long dimension parallel to the long dimension of the original 20.3 × 25.4 cm(2) sheets, were exposed at a depth of 5 cm on a Varian Trilogy at the center of a 20 × 20 cm(2) open field at seven doses between 50 and 1600 cGy using 6 MV photons. These films together with an unexposed film from the same production lot were lined one next to the other on an Epson 10000 XL or 11000 XL scanner in portrait orientation with their long dimension parallel to the scan direction. Scanned images were then obtained with the line of films positioned at seven discrete lateral locations perpendicular to the scan direction. The process was repeated in landscape orientation and on three other Epson scanners. Data were also collected for three additional production lots of EBT3 film (11051302, 03031401, and 03171403). From measurements at the various lateral positions, the scanner response was determined as a function of the lateral position of the scanned film. For a given color channel X, the response at any lateral position L is related to the response at the center, C, of the scanner by Response(C, D, X) = A(L,X) + B(L,X) ⋅ Response(L, D, X), where D is dose and the coefficients A(L,X) and B(L,X) are determined from the film measurements at the center of the scanner and six other discrete lateral positions. The values at intermediate lateral positions were obtained by linear interpolation. The coefficients were determined for the red, green, and blue color channels, preserving the ability to apply triple-channel dosimetry once corrections were applied to compensate for the

  16. Comparison of film measurements and Monte Carlo simulations of dose delivered with very high-energy electron beams in a polystyrene phantom

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Bazalova-Carter, Magdalena; Liu, Michael; Palma, Bianey

    2015-04-15

    Purpose: To measure radiation dose in a water-equivalent medium from very high-energy electron (VHEE) beams and make comparisons to Monte Carlo (MC) simulation results. Methods: Dose in a polystyrene phantom delivered by an experimental VHEE beam line was measured with Gafchromic films for three 50 MeV and two 70 MeV Gaussian beams of 4.0–6.9 mm FWHM and compared to corresponding MC-simulated dose distributions. MC dose in the polystyrene phantom was calculated with the EGSnrc/BEAMnrc and DOSXYZnrc codes based on the experimental setup. Additionally, the effect of 2% beam energy measurement uncertainty and possible non-zero beam angular spread on MC dosemore » distributions was evaluated. Results: MC simulated percentage depth dose (PDD) curves agreed with measurements within 4% for all beam sizes at both 50 and 70 MeV VHEE beams. Central axis PDD at 8 cm depth ranged from 14% to 19% for the 5.4–6.9 mm 50 MeV beams and it ranged from 14% to 18% for the 4.0–4.5 mm 70 MeV beams. MC simulated relative beam profiles of regularly shaped Gaussian beams evaluated at depths of 0.64 to 7.46 cm agreed with measurements to within 5%. A 2% beam energy uncertainty and 0.286° beam angular spread corresponded to a maximum 3.0% and 3.8% difference in depth dose curves of the 50 and 70 MeV electron beams, respectively. Absolute dose differences between MC simulations and film measurements of regularly shaped Gaussian beams were between 10% and 42%. Conclusions: The authors demonstrate that relative dose distributions for VHEE beams of 50–70 MeV can be measured with Gafchromic films and modeled with Monte Carlo simulations to an accuracy of 5%. The reported absolute dose differences likely caused by imperfect beam steering and subsequent charge loss revealed the importance of accurate VHEE beam control and diagnostics.« less

  17. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jordan, K; University of Western Ontario, London, ON, CA

    Purpose: Film densitometers optimized for radiochromic EBT3 film are not commercially available and flatbed document scanners are limited by broad spectral overlap of the red green and blue channels and polarization effects. Non-polarized, narrow bandwith, intensity stabilized light emitting diode (LED) light sources were constructed and transmission measurements compared for point, slot and full aperture geometries. Methods: LED’s with 10 nm bandwidths were focused to 1 mm spots for point scanning. The LED’s back-illuminated a 6 mm diameter diffuser near the focal point of a 25 cm diameter Fresnel lens for slot and full aperture transmission imaging. Films were locatedmore » at the aperture plane, 15 cm from the lens and imaged with a 16 bit digital camera with 50 mm lens. Sheets, 10×8 cm, were irradiated to 25 Gy with a 4×4 cm 6 MV photon beam. Transmission measurements with a 633nm, linearly polarized laser were used as lowacceptance angle reference geometry. Results: LED intensity stabilized to <1% within one minute of startup. The useful transmission range increased as the acceptance angle decreased, laser(∼2%)> spot(∼3%)> slot(∼5%)> full aperture(∼10%). The Fresnel lens introduced a 4-fold polarization symmetry that increased radially causing a 2% change in transmission at 10 cm from the optic axis for EBT3 film. Polarized spot densitometer and laser transmission were similar. Conclusion: Spectrally filtered LED sources were effective for transmission measurements of radiochromic films. Because of polarization sensitivity of EBT3 film, the scanning spot geometry is the most general for measuring unpolarised transmission of larger film sizes, ie > 15×15 cm. Due to EBT3 film scatter, spot scanning should provide the most accurate profiles of high dose gradients, for a given acceptance angle. The spot scanning densitometer provided useful transmission to ∼5, 25, >25 Gy with 635, 590 and 530 nm light respectively.« less

  18. SU-E-T-162: Characterization of a New Oblong Cone for Use with the Intraoperative Mobetron Unit

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Cantley, J; Colussi, V

    Purpose: To evaluate the dosimetric characteristics of a new flat 7 cm x 12 cm oblong cone for intraoperative radiation therapy (IORT) procedures. Methods: Percent depth dose (PDD) curves and profile measurements at dose depths of D100, D90, D80, D50, and D30 were measured using a parallel-plate ion chamber and a 3-D water tank Blue phantom II (iba) for the three energies (6, 9, and 12 MeV) available with the new generation Mobetron 2000. In addition, PDD curves and profiles were made with Gafchromic EBT3 films in solid water phantom. H&D curves were created for each energy for absolute calibration.more » The films were analyzed using the RIT image analysis software and then compared with the ion chamber results. Output values were obtained by normalizing the dose/MU at D100 at the clinical axis of the oblong cone by the dose/MU at D100 for the flat 10cm cylindrical standard cone using ion chamber measurements. Results: Ion chamber results show that the PDDs from the flat oblong cone, at the clinical axis, match with the PDDs for its “cylindrical brother” 7cm flat cone. These PDDs are comparable with those from the 10cm standard cone. The D100 coincides for all three energies. The cone output factors for the oblong cone are 1.07 for all energies. Ion chamber profiles measurements and film analysis show that the use of flat oblong cone with the 6 and 9 MeV energies Result in horns of approximately 12% and 8%, respectively. Conversely, 12 MeV profiles show loss of flatness near the field edge. Conclusion: The potential advantage of the oblong cones is the elimination of the well-known difficulties of electron field matching, both dosimetrically and in clinical setup.« less

  19. Ex vivo validation of a stoichiometric dual energy CT proton stopping power ratio calibration

    NASA Astrophysics Data System (ADS)

    Xie, Yunhe; Ainsley, Christopher; Yin, Lingshu; Zou, Wei; McDonough, James; Solberg, Timothy D.; Lin, Alexander; Teo, Boon-Keng Kevin

    2018-03-01

    A major source of uncertainty in proton therapy is the conversion of Hounsfield unit (HU) to proton stopping power ratio relative to water (SPR). In this study, we measured and quantified the accuracy of a stoichiometric dual energy CT (DECT) SPR calibration. We applied a stoichiometric DECT calibration method to derive the SPR using CT images acquired sequentially at 80 kVp and 140 kVp . The dual energy index was derived based on the HUs of the paired spectral images and used to calculate the effective atomic number (Z eff), relative electron density ({{ρ }e} ), and SPRs of phantom and biological materials. Two methods were used to verify the derived SPRs. The first method measured the sample’s water equivalent thicknesses to deduce the SPRs using a multi-layer ion chamber (MLIC) device. The second method utilized Gafchromic EBT3 film to directly compare relative ranges between sample and water after proton pencil beam irradiation. Ex vivo validation was performed using five different types of frozen animal tissues with the MLIC and three types of fresh animal tissues using film. In addition, the residual ranges recorded on the film were used to compare with those from the treatment planning system using both DECT and SECT derived SPRs. Bland-Altman analysis indicates that the differences between DECT and SPR measurement of tissue surrogates, frozen and fresh animal tissues has a mean of 0.07% and standard deviation of 0.58% compared to 0.55% and 1.94% respectively for single energy CT (SECT) and SPR measurement. Our ex vivo study indicates that the stoichiometric DECT SPR calibration method has the potential to be more accurate than SECT calibration under ideal conditions although beam hardening effects and other image artifacts may increase this uncertainty.

  20. SU-E-T-296: Dosimetric Analysis of Small Animal Image-Guided Irradiator Using High Resolution Optical CT Imaging of 3D Dosimeters

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Na, Y; Qian, X; Wuu, C

    Purpose: To verify the dosimetric characteristics of a small animal image-guided irradiator using a high-resolution of optical CT imaging of 3D dosimeters. Methods: PRESAEGE 3D dosimeters were used to determine dosimetric characteristics of a small animal image-guided irradiator and compared with EBT2 films. Cylindrical PRESAGE dosimeters with 7cm height and 6cm diameter were placed along the central axis of the beam. The films were positioned between 6×6cm{sup 2} cubed plastic water phantoms perpendicular to the beam direction with multiple depths. PRESAGE dosimeters and EBT2 films were then irradiated with the irradiator beams at 220kVp and 13mA. Each of irradiated PRESAGEmore » dosimeters named PA1, PA2, PB1, and PB2, was independently scanned using a high-resolution single laser beam optical CT scanner. The transverse images were reconstructed with a 0.1mm high-resolution pixel. A commercial Epson Expression 10000XL flatbed scanner was used for readout of irradiated EBT2 films at a 0.4mm pixel resolution. PDD curves and beam profiles were measured for the irradiated PRESAGE dosimeters and EBT2 films. Results: The PDD agreements between the irradiated PRESAGE dosimeter PA1, PA2, PB1, PB2 and the EB2 films were 1.7, 2.3, 1.9, and 1.9% for the multiple depths at 1, 5, 10, 15, 20, 30, 40 and 50mm, respectively. The FWHM measurements for each PRESAEGE dosimeter and film agreed with 0.5, 1.1, 0.4, and 1.7%, respectively, at 30mm depth. Both PDD and FWHM measurements for the PRESAGE dosimeters and the films agreed overall within 2%. The 20%–80% penumbral widths of each PRESAGE dosimeter and the film at a given depth were respectively found to be 0.97, 0.91, 0.79, 0.88, and 0.37mm. Conclusion: Dosimetric characteristics of a small animal image-guided irradiator have been demonstrated with the measurements of PRESAGE dosimeter and EB2 film. With the high resolution and accuracy obtained from this 3D dosimetry system, precise targeting small animal irradiation can

  1. Characterization of MOSkin detector for in vivo skin dose measurement during megavoltage radiotherapy

    PubMed Central

    Jong, Wei Loong; Wong, Jeannie Hsiu Ding; Ng, Kwan Hoong; Ho, Gwo Fuang; Cutajar, Dean L.; Rosenfeld, Anatoly B.

    2014-01-01

    In vivo dosimetry is important during radiotherapy to ensure the accuracy of the dose delivered to the treatment volume. A dosimeter should be characterized based on its application before it is used for in vivo dosimetry. In this study, we characterize a new MOSFET‐based detector, the MOSkin detector, on surface for in vivo skin dosimetry. The advantages of the MOSkin detector are its water equivalent depth of measurement of 0.07 mm, small physical size with submicron dosimetric volume, and the ability to provide real‐time readout. A MOSkin detector was calibrated and the reproducibility, linearity, and response over a large dose range to different threshold voltages were determined. Surface dose on solid water phantom was measured using MOSkin detector and compared with Markus ionization chamber and GAFCHROMIC EBT2 film measurements. Dependence in the response of the MOSkin detector on the surface of solid water phantom was also tested for different (i) source to surface distances (SSDs); (ii) field sizes; (iii) surface dose; (iv) radiation incident angles; and (v) wedges. The MOSkin detector showed excellent reproducibility and linearity for dose range of 50 cGy to 300 cGy. The MOSkin detector showed reliable response to different SSDs, field sizes, surface, radiation incident angles, and wedges. The MOSkin detector is suitable for in vivo skin dosimetry. PACS number: 87.55.Qr PMID:25207573

  2. An open source solution for an in-house built dynamic platform for the validation of stereotactic ablative body radiotherapy for VMAT and IMRT.

    PubMed

    Munoz, Luis; Ziebell, Amy; Morton, Jason; Bhat, Madhava

    2016-12-01

    An in-house solution for the verification of dose delivered to a moving phantom as required for the clinical implementation of lung stereotactic ablative body radiation therapy was developed. The superior-inferior movement required to simulate tumour motion during a normal breathing cycle was achieved via the novel use of an Arduino Uno™, a low-cost open-source microcontroller board connected to a high torque servo motor. Slow CT imaging was used to acquire the image set and a 4D cone beam CT (4D-CBCT) verified the efficacy of contoured margins before treatment on the moving phantom. Treatment fields were delivered to a section of a CIRS™ anthropomorphic phantom. Dose verification to the dynamic phantom with Gafchromic EBT3 film using 3 %-1 mm gamma analysis acceptance criteria registered an absolute dose pass rate for IMRT and VMAT of 98 and 96.6 %, respectively. It was verified that 100 % of the PTV received the prescribed dose of 12 Gy per fraction using the dynamic phantom, and no major discrepancy between planned and measured results due to interplay between multileaf collimator sequences and target motion was observed. This study confirmed that the use of an in-house solution using open source hardware and software with existing quality assurance equipment was appropriate in validating a new treatment technique.

  3. A practical three-dimensional dosimetry system for radiation therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Guo Pengyi; Adamovics, John; Oldham, Mark

    2006-10-15

    There is a pressing need for a practical three-dimensional (3D) dosimetry system, convenient for clinical use, and with the accuracy and resolution to enable comprehensive verification of the complex dose distributions typical of modern radiation therapy. Here we introduce a dosimetry system that can achieve this challenge, consisting of a radiochromic dosimeter (PRESAGE trade mark sign ) and a commercial optical computed tomography (CT) scanning system (OCTOPUS trade mark sign ). PRESAGE trade mark sign is a transparent material with compelling properties for dosimetry, including insensitivity of the dose response to atmospheric exposure, a solid texture negating the need formore » an external container (reducing edge effects), and amenability to accurate optical CT scanning due to radiochromic optical contrast as opposed to light-scattering contrast. An evaluation of the performance and viability of the PRESAGE trade mark sign /OCTOPUS, combination for routine clinical 3D dosimetry is presented. The performance of the two components (scanner and dosimeter) was investigated separately prior to full system test. The optical CT scanner has a spatial resolution of {<=}1 mm, geometric accuracy within 1 mm, and high reconstruction linearity (with a R{sup 2} value of 0.9979 and a standard error of estimation of {approx}1%) relative to independent measurement. The overall performance of the PRESAGE trade mark sign /OCTOPUS system was evaluated with respect to a simple known 3D dose distribution, by comparison with GAFCHROMIC[reg] EBT film and the calculated dose from a commissioned planning system. The 'measured' dose distribution in a cylindrical PRESAGE trade mark sign dosimeter (16 cm diameter and 11 cm height) was determined by optical-CT, using a filtered backprojection reconstruction algorithm. A three-way Gamma map comparison (4% dose difference and 4 mm distance to agreement), between the PRESAGE trade mark sign , EBT and calculated dose distributions, showed

  4. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Pino, R; Therriault-Proulx, F; Yang, J

    Purpose: To perform dose profile and output factor measurements for the Exradin W1 plastic scintillation detector (PSD) for the Gamma Knife Perfexion (GKP) collimators in a Lucy phantom and to compare these values to an Exradin A16 ion chamber, EBT3 radiochromic film and treatment planning system (TPS) data. Methods: We used the Exradin W1 PSD which has a small volume, near-water equivalent sensitive element. It has also been shown to be energy independent. This new detector is manufactured and distributed by Standard Imaging, Inc. Measurements were performed for all three collimators (4 mm, 8 mm and 16 mm) for themore » GKP. The Lucy phantom with the PSD inserted was moved in small steps to acquire profiles in all three directions. EBT3 film was inserted in the Lucy phantom and exposed to a single shot for each collimator. Relative output factors were measured using the three detectors while profiles acquired with the PSD were compared to the ones measured with EBT3 radiochromic film. Results: Measured output factors relative to the largest collimator are as followsCollimator PS EBT3 A1616mm 1.000 1.000 1.0008mm 0.892 0.881 0.8834mm 0.795 0.793 0.727 The nominal (vendor) OFs for GKP are 1.000, 0.900, and 0.814, for collimators 16 mm, 8 mm and 4 mm, respectively. There is excellent agreement between all profiles measured with the PSD and EBT3 as well as with the TPS data provided by the vendor. Conclusion: Output factors measured with the W1 were consistent with the ones measured with EBT3 and A16 ion chamber. Measured profiles are in excellent agreement. The W1 detector seems well suited for beam QA for Gamma Knife due to its dosimetric characteristics. Sam Beddar would like to disclose a NIH/NCI SBIR Phase II grant (2R44CA153824-02A1) with Standard Imaging, Title: “Water-Equivalent Plastic Scintillation Detectors for Small Field Radiotherapy”.« less

  5. TU-F-201-01: General Aspects of Radiochromic Film Dosimetry

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Niroomand-Rad, A.

    Since the introduction of radiochromic films (RCF) for radiation dosimetry, the scope of RCF dosimetry has expanded steadily to include many medical applications, such as radiation therapy and diagnostic radiology. The AAPM Task Group (TG) 55 published a report on the recommendations for RCF dosimetry in 1998. As the technology is advancing rapidly, and its routine clinical use is expanding, TG 235 has been formed to provide an update to TG-55 on radiochromic film dosimetry. RCF dosimetry applications in clinical radiotherapy have become even more widespread, expanding from primarily brachytherapy and radiosurgery applications, and gravitating towards (but not limited to)more » external beam therapy (photon, electron and protons), such as quality assurance for IMRT, VMAT, Tomotherapy, SRS/SRT, and SBRT. In addition, RCF applications now extend to measurements of radiation dose in particle beams and patients undergoing medical exams, especially fluoroscopically guided interventional procedures and CT. The densitometers/scanners used for RCF dosimetry have also evolved from the He-Ne laser scanner to CCD-based scanners, including roller-based scanner, light box-based digital camera, and flatbed color scanner. More recently, multichannel RCF dosimetry introduced a new paradigm for external beam dose QA for its high accuracy and efficiency. This course covers in detail the recent advancements in RCF dosimetry. Learning Objectives: Introduce the paradigm shift on multichannel film dosimetry Outline the procedures to achieve accurate dosimetry with a RCF dosimetry system Provide comprehensive guidelines on RCF dosimetry for various clinical applications One of the speakers has a research agreement from Ashland Inc., the manufacturer of Gafchromic film.« less

  6. The effect of Mepitel Film on acute radiation-induced skin reactions in head and neck cancer patients: a feasibility study.

    PubMed

    Wooding, Hayley; Yan, Jing; Yuan, Ling; Chyou, Te-Yu; Gao, Shanbao; Ward, Iain; Herst, Patries M

    2018-01-01

    Mepitel Film significantly decreases acute radiation-induced skin reactions in breast cancer patients. Here we investigated the feasibility of using Mepitel Film in head and neck cancer patients (ACTRN12614000932662). Out of a total of 36 head and neck cancer patients from New Zealand (NZ) (n = 24) and China (n = 12) recruited between June 2015 and December 2016, 33 patients complied with protocol. Of these, 11 NZ patients followed a management protocol; 11 NZ patients and 11 Chinese patients followed a prophylactic protocol. An area of the neck receiving a homogenous radiation dose of > 35 Gy was divided into two equal halves; one half was randomized to Film and the other to either Sorbolene cream (NZ) or Biafine cream (China). Skin reaction severity was measured by Radiation Induced Skin Reaction Assessment Scale and expanded Radiation Therapy Oncology Group toxicity criteria. Skin dose was measured by thermoluminescent dosimeters or gafchromic film. Film decreased overall skin reaction severity (combined Radiation Induced Skin Reaction Assessment Scale score) by 29% and moist desquamation rates by 37% in the Chinese cohort and by 27 and 28%, respectively in the NZ cohort. Mepitel Film did not affect head movements but did not adhere well to the skin, particularly in males with heavy beard stubble, and caused itchiness, particularly in Chinese patients. Mepitel Film reduced acute radiation-induced skin reactions in our head and neck cancer patients, particularly in patients without heavy stubble. Advances in knowledge: This is the first study to confirm the feasibility of using Mepitel Film in head and neck cancer patients.

  7. Micrometer-resolved film dosimetry using a microscope in microbeam radiation therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Bartzsch, Stefan, E-mail: stefan.bartzsch@icr.ac.uk; Oelfke, Uwe; Lott, Johanna

    2015-07-15

    Purpose: Microbeam radiation therapy (MRT) is a still preclinical tumor therapy approach that uses arrays of a few tens of micrometer wide parallel beams separated by a few 100 μm. The production, measurement, and planning of such radiation fields are a challenge up to now. Here, the authors investigate the feasibility of radiochromic film dosimetry in combination with a microscopic readout as a tool to validate peak and valley doses in MRT, which is an important requirement for a future clinical application of the therapy. Methods: Gafchromic{sup ®} HD-810 and HD-V2 films are exposed to MRT fields at the biomedicalmore » beamline ID17 of the European Synchrotron Radiation Facility (ESRF) and are afterward scanned with a microscope. The measured dose is compared with Monte Carlo calculations. Image analysis tools and film handling protocols are developed that allow accurate and reproducible dosimetry. The performance of HD-810 and HD-V2 films is compared and a detailed analysis of the resolution, noise, and energy dependence is carried out. Measurement uncertainties are identified and analyzed. Results: The dose was measured with a resolution of 5 × 1000 μm{sup 2} and an accuracy of 5% in the peak and between 10% and 15% in the valley region. As main causes for dosimetry uncertainties, statistical noise, film inhomogeneities, and calibration errors were identified. Calibration errors strongly increase at low doses and exceeded 3% for doses below 50 and 70 Gy for HD-V2 and HD-810 films, respectively. While the grain size of both film types is approximately 2 μm, the statistical noise in HD-V2 is much higher than in HD-810 films. However, HD-810 films show a higher energy dependence at low photon energies. Conclusions: Both film types are appropriate for dosimetry in MRT and the microscope is superior to the microdensitometer used before at the ESRF with respect to resolution and reproducibility. However, a very careful analysis of the image data is required

  8. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Schnell, E; Ferreira, C; Ahmad, S

    Purpose: Accuracy of a RSP-HU calibration curve produced for proton treatment planning is tested by comparing the treatment planning system dose grid to physical doses delivered on film by a Mevion S250 double-scattering proton unit. Methods: A single batch of EBT3 Gafchromic film was used for calibration and measurements. The film calibration curve was obtained using Mevion proton beam reference option 20 (15cm range, 10cm modulation). Paired films were positioned at the center of the spread out Bragg peak (SOBP) in solid water. The calibration doses were verified with an ion chamber, including background and doses from 20cGy to 350cGy.more » Films were scanned in a flatbed Epson-Expression 10000-XL scanner, and analyzed using the red channel. A Rando phantom was scanned with a GE LightSpeed CT Simulator. A single-field proton plan (Eclipse, Varian) was calculated to deliver 171cGy to the pelvis section (heterogeneous region), using a standard 4×4cm aperture without compensator, 7.89cm beam range, and 5.36cm SOBP. Varied depths of the calculated distal 90% isodose-line were recorded and compared. The dose distribution from film irradiated between Rando slices was compared with the calculated plans using RIT v.6.2. Results: Distal 90% isodose-line depth variation between CT scans was 2mm on average, and 4mm at maximum. Fine calculation of this variation was restricted by the dose calculation grid, as well as the slice thickness. Dose differences between calibrated film measurements and calculated doses were on average 5.93cGy (3.5%), with the large majority of differences forming a normal distribution around 3.5cGy (2%). Calculated doses were almost entirely greater than those measured. Conclusion: RSP to HU calibration curve is shown to produce distal depth variation within the margin of tolerance (±4.3mm) across all potential scan energies and protocols. Dose distribution calculation is accurate to 2–4% within the SOBP, including areas of high tissue

  9. Preliminary study for small animal preclinical hadrontherapy facility

    NASA Astrophysics Data System (ADS)

    Russo, G.; Pisciotta, P.; Cirrone, G. A. P.; Romano, F.; Cammarata, F.; Marchese, V.; Forte, G. I.; Lamia, D.; Minafra, L.; Bravatá, V.; Acquaviva, R.; Gilardi, M. C.; Cuttone, G.

    2017-02-01

    Aim of this work is the study of the preliminary steps to perform a particle treatment of cancer cells inoculated in small animals and to realize a preclinical hadrontherapy facility. A well-defined dosimetric protocol was developed to explicate the steps needed in order to perform a precise proton irradiation in small animals and achieve a highly conformal dose into the target. A precise homemade positioning and holding system for small animals was designed and developed at INFN-LNS in Catania (Italy), where an accurate Monte Carlo simulation was developed, using Geant4 code to simulate the treatment in order to choose the best animal position and perform accurately all the necessary dosimetric evaluations. The Geant4 application can also be used to realize dosimetric studies and its peculiarity consists in the possibility to introduce the real target composition in the simulation using the DICOM micro-CT image. This application was fully validated comparing the results with the experimental measurements. The latter ones were performed at the CATANA (Centro di AdroTerapia e Applicazioni Nucleari Avanzate) facility at INFN-LNS by irradiating both PMMA and water solid phantom. Dosimetric measurements were performed using previously calibrated EBT3 Gafchromic films as a detector and the results were compared with the Geant4 simulation ones. In particular, two different types of dosimetric studies were performed: the first one involved irradiation of a phantom made up of water solid slabs where a layer of EBT3 was alternated with two different slabs in a sandwich configuration, in order to validate the dosimetric distribution. The second one involved irradiation of a PMMA phantom made up of a half hemisphere and some PMMA slabs in order to simulate a subcutaneous tumour configuration, normally used in preclinical studies. In order to evaluate the accordance between experimental and simulation results, two different statistical tests were made: Kolmogorov test and

  10. MO-D-213-04: The Proximity to the Skin of PTV Affects PTV Coverage and Skin Dose for TomoTherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Reynolds, T; Higgins, P; Watanabe, Y

    Purpose: The proximity to the skin surface of the PTV for the patients with skin disease could be a concern in terms of the PTV coverage and actual surface dose (SD). IMRT optimization algorithms increase the beam intensity close to the skin in order to compensate for lack of scattering material, leading to enhanced SD but potential hot spots. This study aims to investigate the effect of PTV proximity to the skin on planning and measured SD Methods: All measurements were done for 6 MV X-ray beam of Helical TomoTherapy. An anthropomorphic phantom was scanned in a CT simulator inmore » a routine manner with thermoplastic mask immobilization. PTVs were created with varying distances to the skin of 0 mm -(PTV1), 1 mm- (PTV2), 2 mm-(PTV3) and 3 mm-(PTV4). Also, a 5 mm bolus was used with PTV1 (PTV5). All planning constraints were kept the same in all studies (hard constraint: 95% of the prescription dose covered 95% of the PTV). Gafchromic film (EBT3) was placed under the mask on the phantom surface, and the resulting dose was estimated using RIT software. Results: Optimizing the dose using different PTVs lead to average planned target doses of 10.8, 10.3, 10.2, 10.3 and 10.0 Gy, with maximum doses 12.2, 11.2, 11.1, 11.1 and 10.0 Gy for PTV1, PTV2, PTV3, PTV4 and PTV5, respectively. EBT3 measurements indicated a significant decrease of SD with skin distance by 12.7% (PTV1), 21.9% (PTV2), 24.8% (PTV3) and 28.4% (PTV4) comparing to prescription dose. Placement of a 5 mm bolus on the phantom surface resulted in a SD close to prescribed (+0.5%). Conclusion: This work provides a clear demonstration of the relationship between the skin dose and the PTV to the skin distance. The results indicate the necessity of a bolus even for TomoTherapy when high skin dose is required.« less

  11. Building an Evidence Base for Effective Supervision Practices: An Analogue Experiment of Supervision to Increase EBT Fidelity.

    PubMed

    Bearman, Sarah Kate; Schneiderman, Robyn L; Zoloth, Emma

    2017-03-01

    Treatments that are efficacious in research trials perform less well under routine conditions; differences in supervision may be one contributing factor. This study compared the effect of supervision using active learning techniques (e.g. role play, corrective feedback) versus "supervision as usual" on therapist cognitive restructuring fidelity, overall CBT competence, and CBT expertise. Forty therapist trainees attended a training workshop and were randomized to supervision condition. Outcomes were assessed using behavioral rehearsals pre- and immediately post-training, and after three supervision meetings. EBT knowledge, attitudes, and fidelity improved for all participants post-training, but only the SUP+ group demonstrated improvement following supervision.

  12. Dosimetry Formalism and Implementation of a Homogenous Irradiation Protocol to Improve the Accuracy of Small Animal Whole-Body Irradiation Using a Cesium-137 Irradiator

    PubMed Central

    Brodin, N. Patrik; Chen, Yong; Yaparpalvi, Ravindra; Guha, Chandan; Tomé, Wolfgang A.

    2015-01-01

    Shielded 137Cs irradiators are routinely used in pre-clinical radiation research to perform in vitro or in vivo investigations. Without appropriate dosimetry and irradiation protocols in place, there can be large uncertainty in the delivered dose of radiation between irradiated subjects that could lead to inaccurate and possibly misleading results. Here, a dosimetric evaluation of the JL Shepard Mark I-68A 137Cs irradiator and an irradiation technique for whole-body irradiation of small animals that allows one to limit the between subject variation in delivered dose to ±3% are provided. Mathematical simulation techniques and Gafchromic EBT film were used to describe the region within the irradiation cavity with homogeneous dose distribution (100% ±5%), the dosimetric impact of varying source-to-subject distance, and the variation in attenuation thickness due to turntable rotation. Furthermore, an irradiation protocol and dosimetry formalism that allows calculation of irradiation time for whole-body irradiation of small animals is proposed, that is designed to ensure a more consistent dose delivery between irradiated subjects. To compare this protocol with the conventional irradiation protocol suggested by the vendor, high-resolution film dosimetry measurements evaluating the dose difference between irradiation subjects and the dose distribution throughout subjects was performed, using phantoms resembling small animals. Based on these results, there can be considerable variation in the delivered dose of > ±5% using the conventional irradiation protocol for whole-body irradiation doses below 5 Gy. Using the proposed irradiation protocol this variability can be reduced to within ±3% and the dosimetry formalism allows for more accurate calculation of the irradiation time in relation to the intended prescription dose. PMID:26710162

  13. [Prostate radiation therapy: in vivo measurement of the dose delivered by kV-CBCT].

    PubMed

    Marinello, G; Mege, J-P; Besse, M-C; Kerneur, G; Lagrange, J-L

    2009-09-01

    To investigate if the regular use of kV-CBCT notably increases the dose delivered to tumor and surrounding healthy tissues. Images were obtained using a Varian equipment (OBI version 1.3, 645 to 650 projections in 370 degrees to acquire image), and patients were irradiated at source-tumor distance: 100cm. In vivo measurements were performed using radio-thermoluminescent dosimeters Harshaw-TLD700H (TLD) at skin (anterior-posterior and lateral axis crossing the rotation axis), with a fourth TLD group under the table thanks to a retrolaser. TLD's were calibrated at the kV-CBCT effective energy (64 keV), and the method validated using an anthropomorphic phantom, in which Gafchromic EBT films were also inserted. The phantom study showed that the dose distribution depends on the phantom position relative to the axis and that the doses measured at the phantom surface using TLD and films (good agreement) were maximum at the entrance of the anterior-posterior axis. Their arithmetic mean was equal, or a slightly greater than doses measured at mid-thickness of the phantom and at the level of the rectum (OAR). In vivo measurements performed on the five first patients (125 kV-CBCT) yield a mean skin dose per kV-CBCT varying from 5.8+/-0.1 to 7.3+/-0.2 cGy on the anterior-posterior axis. Lateral skin doses vary from 3.4+/-0.2 to 4.5+/-0.2 cGy. Doses delivered by repeated kV-CBCT are not negligible. They should be taken into account, but questions about the RBE to be applied to kilovoltage X-rays are raised.

  14. Volumetric Modulated Arc Radiotherapy for Vestibular Schwannomas

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lagerwaard, Frank J.; Meijer, Otto W.M.; Hoorn, Elles A.P. van der

    2009-06-01

    Purpose: To evaluate volumetric modulated arc radiotherapy (RapidArc [RA]), a novel approach allowing for rapid treatment delivery, for the treatment of vestibular schwannoma (VS). Methods and Materials: The RA plans were generated for a small (0.5 cm{sup 3}), intermediate (2.8 cm{sup 3}), and large (14.8 cm{sup 3}) VS. The prescription dose was 12.5 Gy to the encompassing 80% isodose. The RA plans were compared with conventional radiosurgery plans using both a single dynamic conformal arc (1DCA) and five noncoplanar dynamic conformal arcs (5DCA). Conformity indices (CI) and dose-volume histograms of critical organs were compared. The RA plan for the medium-sizedmore » VS was measured in a phantom using Gafchromic EBT films and compared with calculated dose distributions. Results: The RA planning was completed within 30 min in all cases, and calculated treatment delivery time (after patient setup) was 5 min vs. 20 min for 5DCA. A superior CI was achieved with RA, with a substantial decrease in low-dose irradiation of the normal brain achieved relative to 5DCA plans. Maximum doses to critical organs were similar for RA and 5DCA but were higher for 1DCA. Film measurements showed the differences between calculated and measured doses to be smaller than 1.5% in the high-dose area and smaller than 3% in the low-dose area. Conclusion: The RA plans consistently achieved a higher CI and decrease in areas of low-dose irradiation. This, together with shorter treatment delivery times, has led to RA replacing our conventional five-arc radiosurgery technique for VS.« less

  15. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Xu, Z; Baker, J; Hsia, A

    Purpose: The commercially available Leipzig-style Cone for High Dose Rate (HDR) Brachytherapy has a steep depth dose curve and a non-uniform dose distribution. This work shows the performance of a Ring Surface Applicator created using a 3D printer that can generate a better dose distribution. Calculated doses were verified with film measurement. Methods: The water equivalent red-ABS plastic was used to print the Ring Surface Applicator which hosts three catheters: a center piece with a straight catheter and two concentric rings with diameters of 3.5 and 5.5 cm. Gafchromic EBT2 film, Epson Expression 10000 flatbed scanner, and the online softwaremore » at radiochromic.com were used to analyze the measured data. 10cm×10cm piece of film was sandwiched between two 15×10×5cm3 polystyrene phantoms. The applicator was positioned directly on top of the phantom. Measurement was done using dwell time and positions calculated by Eclipse BrachyVision treatment planning system (RTP). Results: Depth dose curve was generated from the plan and measurement. The results show that the measured and calculated depth dose were in agreement (<3%) from surface to 4mm depth. A discrepancy of 6% was observed at 5 mm depth, where the dose is typically prescribed to. For depths deeper than 5 mm, the measured doses were lower than those calculated by Eclipse BrachyVision. This can be attributed to a combination of simple calculation algorithm using TG-43 and the lack of inhomogeneity correction. Dose profiles at 5 mm depth were also generated from TPS calculation and measured with film. The measured and calculated profiles are similar. Consistent with the depth dose curve, the measured dose is lower than the calculated. Conclusion: Our results showed that the Ring Surface Applicator, printed using 3D printer, can generate more uniform dose distribution within the target volume and can be safely used in the clinic.« less

  16. SU-G-201-07: Dosimetric Verification of a 3D Printed HDR Skin Brachytherapy Applicator

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rasmussen, K; Stanley, D; Eng, T

    Purpose: The use of radiation as a treatment modality for skin cancer has increased significantly over the last decade with standardized applicators. Utilizing 3D printing, the ability to make applicators specifically designed for each patient’s anatomy has become economically feasible. With this in mind it was the aim of this study to determine the dosimetric accuracy of a 3-D printed HDR brachytherapy applicator for the skin. Methods: A CT reference image was used to generate a custom applicator based on an anthropomorphic head and neck phantom. To create the applicator a 1cm expansion anteriorly with 0.5cmX0.5cm trenches on the outermore » surface that were spaced 1cm sup-inf to accommodate standard 6F flexible catheters. The applicator was printed using PLA material using a printrbot simple printer. A treatment plan optimized to deliver a clinically representative volume was created in Oncentra and delivered with a nucletron afterloader. Measurements were made using TLDs and EBT3 gafchromic film that were placed between the applicator and the phantom’s forehead. An additional piece of film was also used to qualitatively asses the dose distribution in the transverse plane. Using a standard vaginal cylinder and bolus, a standardized curve correlating TLD and film exposure-to-radiation dose was established by irradiating film to known doses (200,500,700 cGy) at a 3.5 cm radius distance. Results: Evaluated TLDs showed the absolute dose delivered to the skin surface using the 3-D printed bolus was 615cGy±6%, with a mean predicted TPS value in the measured area of 617.5±7%. Additionally, planar dose distributions had good qualitative agreement with calculated TPS isodoses. Conclusion: This work demonstrates patient specific 3-D printed HDR brachytherapy applicators for skin cancer treatments are practical and accurate in TPS calculations but additional measurements are needed to verify additional sites and dose at depth.« less

  17. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Li, K; Able, A

    Purpose: To evaluate an Enhanced Dynamic Wedge (EDW) as part of machine commission process with feature study. Methods: The EDW system in this study was from a Truebeam, which is the Linear accelerator manufactured by Varian Medical Systems. The EDW feature vectors includes selected elements. These elements were dosimetric output spots check, field size, wedge angles, dose rate, collimator orientation, and different energy settings. Point dose measurement was done by a PTW farmer chamber, and profiles were measured by Gafchromic EBT2 films positing at different depths of the Solidwater based on the study elements. The output spot measurements were donemore » with PTW farmer chamber with Solidwater setting for all orientation and wedge angles in the EDW system. The profiles comparisons were done by IMRT measurement function in RIT software at version 6.3. And the films were scanned by Vidar scanner. Dosimetry calculation were done by using the same Solidwater scanned by GE LightSpeed CT in Eclipse Treatment Planning System (TPS). Then measurements were compared to simulation results in TPS. Results: The energy average percentage difference between chamber measurement and TPS was 0.16% with standard deviation (SD) at 0.93%. For selected features, the average percentage difference between film measurement and computation was 0.93% with SD at 1.55% in horizontal profiles, and 1.18% with SD at 0.98% at vertical profiles. The average gamma difference for film measurement and TPS computing results was at 0.924 with SD at 0.314. Conclusion: A feature vector was developed to describe the commission of EDW, and developing a complete set of features for sufficiency of commission of a LINAC function could provide optimal commission instance with acceptable confident level of clinical application of the machine. Given the institution specific vector pattern and big data process, it could provide wide range clinical outcome comparison information in application of EDW.« less

  18. Feasibility study of proton-based quality assurance of proton range compensator

    NASA Astrophysics Data System (ADS)

    Park, S.; Jeong, C.; Min, B. J.; Kwak, J.; Lee, J.; Cho, S.; Shin, D.; Lim, Y. K.; Park, S. Y.; Lee, S. B.

    2013-06-01

    All patient specific range compensators (RCs) are customized for achieving distal dose conformity of target volume in passively scattered proton therapy. Compensators are milled precisely using a computerized machine. In proton therapy, precision of the compensator is critical and quality assurance (QA) is required to protect normal tissues and organs from radiation damage. This study aims to evaluate the precision of proton-based quality assurance of range compensator. First, the geometry information of two compensators was extracted from the DICOM Radiotherapy (RT) plan. Next, RCs were irradiated on the EBT film individually by proton beam which is modulated to have a photon-like percent depth dose (PDD). Step phantoms were also irradiated on the EBT film to generate calibration curve which indicates relationship between optical density of irradiated film and perpendicular depth of compensator. Comparisons were made using the mean absolute difference (MAD) between coordinate information from DICOM RT and converted depth information from the EBT film. MAD over the whole region was 1.7, and 2.0 mm. However, MAD over the relatively flat regions on each compensator selected for comparison was within 1 mm. These results shows that proton-based quality assurance of range compensator is feasible and it is expected to achieve MAD over the whole region less than 1 mm with further correction about scattering effect of proton imaging.

  19. SU-F-T-550: Radiochromic Plastic Thin Sheet Dosimeter: Initial Performance

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jordan, K; Adamovics, J

    Purpose: Thin sheets, of a high sensitivity formulation of radiochromic dosimeter, Presage were prepared and evaluated for optical readout. Methods: Sheets of radiochromic polyurethane, 12 cm long, 10 cm wide and 0.2 cm thick were prepared with leuco crystal violet as the reporter molecule. Sample transmission was evaluated at a wavelength of 590 nm with in-house constructed instruments: optical cone beam laser CT scanner, fixed and scanning spot densitometers. Sample sequential irradiations to a total dose of 40 Gy were conducted with a modified, Theratron 60, cobalt radiotherapy machine at dose rates of 1 or 0.25 Gy per minute. Exposuremore » to ambient and readout light was minimized to limit background photochromic signals. Samples were stored at 4°C. Optical activity was assessed from linearly polarized transmission images. Comparison sensitivity measurements with EBT3 film were conducted. Results: Samples were transparent, smooth and pale purple before irradiation. Radiochromic reaction was completed in less than 5 minutes. A linear dose response with a sensitivity of 0.5 cm-1Gy-1 was observed. Micrometer measurements found sheet thickness variations up to 20%. Uniform dose, 2 Gy attenuation images, correlated with local sheet thicknesses. Comparable measurements with EBT3 film were 3 times more sensitive at 1 Gy but above 15 Gy, EBT3 film had lower sensitivity than 0.2 cm thick Presage sheet dosimeter due to its non-linear response. Conclusion: Dose sensitivity provided a 10% decrease in transmission for a 1 Gy dose. Improvements in mold design are expected to allow production of sheets with less than 5% variation in thickness. Above, 10 Gy, Presage sheet dosimeter performance expected to exceed EBT3 film based on linearity, sensitivity, transparency and smoothness of samples. J Adamovics is owner of Heuris Inc.« less

  20. Accuracy of a dose-area product compared to an absorbed dose to water at a point in a 2 cm diameter field

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Dufreneix, S.; Ostrowsky, A.; Rapp, B.

    Purpose: Graphite calorimeters with a core diameter larger than the beam can be used to establish dosimetric references in small fields. The dose-area product (DAP) measured can theoretically be linked to an absorbed dose at a point by the determination of a profile correction. This study aims at comparing the DAP-based protocol to the usual absorbed dose at a point protocol in a 2 cm diameter field for which both references exist. Methods: Two calorimeters were used, respectively, with a sensitive volume of 0.6 cm (for the absorbed dose at a point measurement) and 3 cm diameter (for the DAPmore » measurement). Profile correction was calculated from a 2D dose mapping using three detectors: a PinPoint chamber, a synthetic diamond, and EBT3 films. A specific protocol to read EBT3 films was implemented and the dose-rate and energy dependences were studied to assure a precise measurement, especially in the penumbra and out-of-field regions. Results: EBT3 films were found independent on dose rates over the range studied but showed a strong under-response (18%) at low energies. Depending on the dosimeter used for calculating the profile correction, a deviation of 0.8% (PinPoint chamber), 0.9% (diamond), or 1.9% (EBT3 films) was observed between the calibration coefficient derived from DAP measurements and the one directly established in terms of absorbed dose to water at a point. Conclusions: The DAP method can currently be linked to the classical dosimetric reference system based in an absorbed dose at a point only with a confidence interval of 95% (k = 2). None of the detectors studied can be used to determine an absorbed dose to water at a point from a DAP measurement with an uncertainty smaller than 1.2%.« less

  1. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rampado, O.; Garelli, E.; Ropolo, R.

    Gafchromic XR-QA films were developed for patient dosimetry in diagnostic radiology. A possible application of these films is the measurement of doses in computed tomography. In this study a method to evaluate the CTDI using Gafchromic XR-QA film and a flatbed scanner was developed and tested. Film samples were cut to dimensions of 6x170 mm{sup 2} in order to have an integration area similar to that of a pencil ionization chamber, with the possibility of changing the integration length. Prior to exposing these films to a computed tomography beam, the angular dependence of the film dose response was investigated bymore » exposing film strips to a static x-ray beam at different angles in the range 0 deg. - 180 deg. A difference of 49% was found between the response with the axis beam parallel to the film surface (90 deg.) and with the axis beam perpendicular (0 deg. and 180 deg.). Integrating over a 360 deg. exposure like the one in computed tomography, a difference of less than 2% was estimated, which is comparable with the measurement error obtainable with XR-QA film. A calibration with a CT beam in the scout mode was performed and film strips were then exposed to single axial scans and to helical scans both in air and in phantoms. Two different types of flatbed scanners were used to read the film samples, a Microtek ScanMaker 9800XL scanner and an Epson Expression 10000 XL scanner, and the accuracy of the results were compared. For beam collimations above 10 mm differences between CTDI measured by film and CTDI measured by ionization chamber below 9% were found for the Epson scanner, with an average estimated error at 1 {sigma} level of 5%. For the Microtek scanner and for the same film samples, differences below 11% with an average error at 1 {sigma} level of 8% were founded. The 1 {sigma} uncertainty of the measured CTDI was provided by the method for each measurement, and it was shown that about the 95% of the differences between the CTDI measurements with

  2. MCNP modelling of vaginal and uterine applicators used in intracavitary brachytherapy and comparison with radiochromic film measurements

    NASA Astrophysics Data System (ADS)

    Ceccolini, E.; Gerardy, I.; Ródenas, J.; van Dycke, M.; Gallardo, S.; Mostacci, D.

    Brachytherapy is an advanced cancer treatment that is minimally invasive, minimising radiation exposure to the surrounding healthy tissues. Microselectron© Nucletron devices with 192Ir source can be used for gynaecological brachytherapy, in patients with vaginal or uterine cancer. Measurements of isodose curves have been performed in a PMMA phantom and compared with Monte Carlo calculations and TPS (Plato software of Nucletron BPS 14.2) evaluation. The isodose measurements have been performed with radiochromic films (Gafchromic EBT©). The dose matrix has been obtained after digitalisation and use of a dose calibration curve obtained with a 6 MV photon beam provided by a medical linear accelerator. A comparison between the calculated and the measured matrix has been performed. The calculated dose matrix is obtained with a simulation using the MCNP5 Monte Carlo code (F4MESH tally).

  3. Characterization and attenuation study on tannin-added Rhizophora spp. particleboard at high energy photon and electron

    NASA Astrophysics Data System (ADS)

    Yusof, Mohd Fahmi Mohd; Hamid, Puteri Nor Khatijah Abd; Tajuddin, Abd Aziz; Abdullah, Reduan; Hashim, Rokiah; Bauk, Sabar; Isa, Norriza Mohd; Isa, Muhammad Jamal Md

    2017-01-01

    The effective atomic number of tannin-added Rhizophora spp. particleboards was determined based on elemental composition using Energy Dispersive X-ray Analysis (EDXA). The value of mass attenuation coefficients were measured using 137Cs and 60Co gamma energies. The attenuation properties of PDD curves and beam profile of tannin-added Rhizophora spp. particleboards were investigated using Gafchromic EBT2 film at 6 MV photon and 6 MeV electrons and compared to the value in water and solid water phantoms. The results showed that tannin-added Rhizophora spp. particleboards having effective atomic number close to the value of water. The mass attenuation coefficients were near to the value of water with χ2 values of 0.018 and 0.357 to 137Cs and 60Co gamma energies respectively. The PDD of tannin-added Rhizophora spp. particleboards at 6 MV photons showed good agreement within 3.21 and 5.91% to that in solid water phantoms and water respectively. The PDD at 6 MeV electrons showed a good agreement within 3.32 and 3.12% to that in solid water phantoms and water respectively. The depth of R50 and R90 in tannin-added Rhizophora spp. also showed a good agreement to that in water and solid water pahtoms. Lower surface dose was observed in tannin-added Rhizophora spp. particleboards at electron beams in comparison to solid water phantoms and water.

  4. The characteristics of dose at mass interface on lung cancer Stereotactic Body Radiotherapy (SBRT) simulation

    NASA Astrophysics Data System (ADS)

    Wulansari, I. H.; Wibowo, W. E.; Pawiro, S. A.

    2017-05-01

    In lung cancer cases, there exists a difficulty for the Treatment Planning System (TPS) to predict the dose at or near the mass interface. This error prediction might influence the minimum or maximum dose received by lung cancer. In addition to target motion, the target dose prediction error also contributes in the combined error during the course of treatment. The objective of this work was to verify dose plan calculated by adaptive convolution algorithm in Pinnacle3 at the mass interface against a set of measurement. The measurement was performed using Gafchromic EBT 3 film in static and dynamic CIRS phantom with amplitudes of 5 mm, 10 mm, and 20 mm in superior-inferior motion direction. Static and dynamic phantom were scanned with fast CT and slow CT before planned. The results showed that adaptive convolution algorithm mostly predicted mass interface dose lower than the measured dose in a range of -0,63% to 8,37% for static phantom in fast CT scanning and -0,27% to 15,9% for static phantom in slow CT scanning. In dynamic phantom, this algorithm was predicted mass interface dose higher than measured dose up to -89% for fast CT and varied from -17% until 37% for slow CT. This interface of dose differences caused the dose mass decreased in fast CT, except for 10 mm motion amplitude, and increased in slow CT for the greater amplitude of motion.

  5. Evaluation and clinical implementation of in vivo dosimetry for kV radiotherapy using radiochromic film and micro-silica bead thermoluminescent detectors.

    PubMed

    Palmer, Antony L; Jafari, Shakardokht M; Mone, Ioanna; Muscat, Sarah

    2017-10-01

    kV radiotherapy treatment calculations are based on flat, homogenous, full-scatter reference conditions. However, clinical treatments often include surface irregularities and inhomogeneities, causing uncertainty. Therefore, confirmation of actual delivered doses in vivo is valuable. The current study evaluates, and implements, radiochromic film and micro silica bead TLD for in vivo kV dosimetry. The kV energy and dose response of EBT3 film and silica bead TLD was established and uncertainty budgets determined. In vivo dosimetry measurements were made for a consecutive series of 30 patients using the two dosimetry systems. Energy dependent calibration factors were required for both dosimetry systems. The standard uncertainty estimate for in vivo measurement with film was 1.7% and for beads was 1.5%. The mean measured dose was -2.1% for film and -2.6% for beads compared to prescription. Deviations up to -9% were found in cases of large surface irregularity, or with underlying air cavities or bone. Dose shielding by beads could be clinically relevant at low kV energies and superficial depths. Both film and beads may be used to provide in vivo verification of delivered doses in kV radiotherapy, particularly for complex situations that are not well represented by standard reference condition calculations. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  6. Monte Carlo simulation and film dosimetry for electron therapy in vicinity of a titanium mesh

    PubMed Central

    Rostampour, Masoumeh; Roayaei, Mahnaz

    2014-01-01

    Titanium (Ti) mesh plates are used as a bone replacement in brain tumor surgeries. In the case of radiotherapy, these plates might interfere with the beam path. The purpose of this study is to evaluate the effect of titanium mesh on the dose distribution of electron fields. Simulations were performed using Monte Carlo BEAMnrc and DOSXYZnrc codes for 6 and 10 MeV electron beams. In Monte Carlo simulation, the shape of the titanium mesh was simulated. The simulated titanium mesh was considered as the one which is used in head and neck surgery with a thickness of 0.055 cm. First, by simulation, the percentage depth dose was obtained while the titanium mesh was present, and these values were then compared with the depth dose of homogeneous phantom with no titanium mesh. In the experimental measurements, the values of depth dose with titanium mesh and without titanium mesh in various depths were measured. The experiments were performed using a RW3 phantom with GAFCHROMIC EBT2 film. The results of experimental measurements were compared with values of depth dose obtained by simulation. In Monte Carlo simulation, as well as experimental measurements, for the voxels immediately beyond the titanium mesh, the change of the dose were evaluated. For this purpose the ratio of the dose for the case with titanium to the case without titanium was calculated as a function of titanium depth. For the voxels before the titanium mesh there was always an increase of the dose up to 13% with respect to the same voxel with no titanium mesh. This is because of the increased back scattering effect of the titanium mesh. The results also showed that for the voxel right beyond the titanium mesh, there is an increased or decreased dose to soft tissues, depending on the depth of the titanium mesh. For the regions before the depth of maximum dose, there is an increase of the dose up to 10% compared to the dose of the same depth in homogeneous phantom. Beyond the depth of maximum dose, there was a

  7. Feasibility of TCP-based dose painting by numbers applied to a prostate case with (18)F-choline PET imaging.

    PubMed

    Dirscherl, Thomas; Rickhey, Mark; Bogner, Ludwig

    2012-02-01

    A biologically adaptive radiation treatment method to maximize the TCP is shown. Functional imaging is used to acquire a heterogeneous dose prescription in terms of Dose Painting by Numbers and to create a patient-specific IMRT plan. Adapted from a method for selective dose escalation under the guidance of spatial biology distribution, a model, which translates heterogeneously distributed radiobiological parameters into voxelwise dose prescriptions, was developed. At the example of a prostate case with (18)F-choline PET imaging, different sets of reported values for the parameters were examined concerning their resulting range of dose values. Furthermore, the influence of each parameter of the linear-quadratic model was investigated. A correlation between PET signal and proliferation as well as cell density was assumed. Using our in-house treatment planning software Direct Monte Carlo Optimization (DMCO), a treatment plan based on the obtained dose prescription was generated. Gafchromic EBT films were irradiated for evaluation. When a TCP of 95% was aimed at, the maximal dose in a voxel of the prescription exceeded 100Gy for most considered parameter sets. One of the parameter sets resulted in a dose range of 87.1Gy to 99.3Gy, yielding a TCP of 94.7%, and was investigated more closely. The TCP of the plan decreased to 73.5% after optimization based on that prescription. The dose difference histogram of optimized and prescribed dose revealed a mean of -1.64Gy and a standard deviation of 4.02Gy. Film verification showed a reasonable agreement of planned and delivered dose. If the distribution of radiobiological parameters within a tumor is known, this model can be used to create a dose-painting by numbers plan which maximizes the TCP. It could be shown, that such a heterogeneous dose distribution is technically feasible. Copyright © 2012. Published by Elsevier GmbH.

  8. SU-E-T-115: Dose Perturbation Study of Self-Expandable Metal and Polyester Esophageal Stents in Proton Therapy Beams

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lee, S; Li, Z; Jalaj, S

    2014-06-01

    Purpose: This work investigates dose perturbations due to Self-expandable metal and polyester esophageal stents undergoing proton radiotherapy for esophageal cancer. Methods: Five commercially available esophageal stents made of nitinol (Evolution, Wallflex and Ultraflex), stainless steel (Z-Stent) and polyester (Polyflex) were tested. Radiochromic film (GafChromic EBT3 film, Ashland, Covington, KY) wrapped around a stent and a 12cc syringe was irradiated with 2CGE (Cobalt Gray Equivalent) of proton beam in a custom fabricated acrylic phantom. An air-hollow syringe simulates the esophagus. Results: The Z-stent created the largest dose perturbations ranges from -14.5% to 6.1% due to the steel composition. The WallFlex, Evolutionmore » and Ultraflex stents produced the dose perturbation ranges of (−9.2%∼8.6%), (−6.8%∼5.7%) and (−6.2%∼6.2%), respectively. The PolyFlex stent contains the radiopaque tungsten markers located top, middle and bottom portions. When the focal cold spots induced by the markers were excluded in the analysis, the dose perturbation range was changed from (−11.6%∼6.4%) to (−0.6%∼5.0%). Conclusion: The magnitude of dose perturbation is related to material of a metallic stent. The non-metallic stent such as PolyFlex shows relatively lower dose perturbation than metallic stents except a radiopaque marker region. Overall Evolution and Ultraflex stent appear to be less dose perturbations. The largest dose perturbations (cold spots) were located at both edges of stents in distal area for the single proton beam irradiation study. The analysis of more than two proton beam which is more typical clinical beam arrangement would be necessary to minimize the doe perturbation effect in proton ratiotherapy.« less

  9. Technical Report: TG-142 compliant and comprehensive quality assurance tests for respiratory gating

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Woods, Kyle; Rong, Yi, E-mail: yrong@ucdavis.edu

    2015-11-15

    Purpose: To develop and establish a comprehensive gating commissioning and quality assurance procedure in compliance with TG-142. Methods: Eight Varian TrueBeam Linacs were used for this study. Gating commissioning included an end-to-end test and baseline establishment. The end-to-end test was performed using a CIRS dynamic thoracic phantom with a moving cylinder inside the lung, which was used for carrying both optically simulated luminescence detectors (OSLDs) and Gafchromic EBT2 films while the target is moving, for a point dose check and 2D profile check. In addition, baselines were established for beam-on temporal delay and calibration of the surrogate, for both megavoltagemore » (MV) and kilovoltage (kV) beams. A motion simulation device (MotionSim) was used to provide periodic motion on a platform, in synchronizing with a surrogate motion. The overall accuracy and uncertainties were analyzed and compared. Results: The OSLD readings were within 5% compared to the planned dose (within measurement uncertainty) for both phase and amplitude gated deliveries. Film results showed less than 3% agreement to the predicted dose with a standard sinusoid motion. The gate-on temporal accuracy was averaged at 139 ± 10 ms for MV beams and 92 ± 11 ms for kV beams. The temporal delay of the surrogate motion depends on the motion speed and was averaged at 54.6 ± 3.1 ms for slow, 24.9 ± 2.9 ms for intermediate, and 23.0 ± 20.1 ms for fast speed. Conclusions: A comprehensive gating commissioning procedure was introduced for verifying the output accuracy and establishing the temporal accuracy baselines with respiratory gating. The baselines are needed for routine quality assurance tests, as suggested by TG-142.« less

  10. Advantage of 3D volumetric dosemeter in delivery quality assurance of dynamic arc therapy: comparison of pencil beam and Monte Carlo calculations

    PubMed Central

    Shin, H-J; Song, J H; Jung, J-Y; Kwak, Y-K; Kay, C S; Kang, Y-N; Choi, B O; Jang, H S

    2013-01-01

    Objective: To evaluate the accuracy of pencil beam calculation (PBC) and Monte Carlo calculation (MCC) for dynamic arc therapy (DAT) in a cylindrically shaped homogenous phantom, by comparing the two plans with an ion chamber, a film and a three-dimensional (3D) volumetric dosemeter. Methods: For this study, an in-house phantom was constructed, and the PBC and MCC plans for DAT were performed using iPlan® RT (BrainLAB®, Heimstetten, Germany). The A16 micro ion chamber (Standard Imaging, Middleton, WI), Gafchromic® EBT2 film (International Specialty Products, Wayne, NJ) and ArcCHECK™ (Sun Nuclear, Melbourne, FL) were used for measurements. For comparison with each plan, two-dimensional (2D) and 3D gamma analyses were performed using 3%/3 mm and 2%/2 mm criteria. Results: The difference between the PBC and MCC plans using 2D and 3D gamma analyses was found to be 7.85% and 28.8%, respectively. The ion chamber and 2D dose distribution measurements did not exhibit this difference revealed by the comparison between the PBC and MCC plans. However, the 3D assessment showed a significant difference between the PBC and MCC (62.7% for PBC vs 93.4% for MCC, p = 0.034). Conclusion: Evaluation using a 3D volumetric dosemeter can be clinically useful for delivery quality assurance (QA), and the MCC should be used to achieve the most reliable dose calculation for DAT. Advances in knowledge: (1) The DAT plan calculated using the PBC has a limitation in the calculation methods, and a 3D volumetric dosemeter was found to be an adequate tool for delivery QA of DAT. (2) The MCC was superior to PBC in terms of the accuracy in dose calculation for DAT even in the homogenous condition. PMID:24234583

  11. Technical Report: TG-142 compliant and comprehensive quality assurance tests for respiratory gating.

    PubMed

    Woods, Kyle; Rong, Yi

    2015-11-01

    To develop and establish a comprehensive gating commissioning and quality assurance procedure in compliance with TG-142. Eight Varian TrueBeam Linacs were used for this study. Gating commissioning included an end-to-end test and baseline establishment. The end-to-end test was performed using a CIRS dynamic thoracic phantom with a moving cylinder inside the lung, which was used for carrying both optically simulated luminescence detectors (OSLDs) and Gafchromic EBT2 films while the target is moving, for a point dose check and 2D profile check. In addition, baselines were established for beam-on temporal delay and calibration of the surrogate, for both megavoltage (MV) and kilovoltage (kV) beams. A motion simulation device (MotionSim) was used to provide periodic motion on a platform, in synchronizing with a surrogate motion. The overall accuracy and uncertainties were analyzed and compared. The OSLD readings were within 5% compared to the planned dose (within measurement uncertainty) for both phase and amplitude gated deliveries. Film results showed less than 3% agreement to the predicted dose with a standard sinusoid motion. The gate-on temporal accuracy was averaged at 139±10 ms for MV beams and 92±11 ms for kV beams. The temporal delay of the surrogate motion depends on the motion speed and was averaged at 54.6±3.1 ms for slow, 24.9±2.9 ms for intermediate, and 23.0±20.1 ms for fast speed. A comprehensive gating commissioning procedure was introduced for verifying the output accuracy and establishing the temporal accuracy baselines with respiratory gating. The baselines are needed for routine quality assurance tests, as suggested by TG-142.

  12. Fast protocol for radiochromic film dosimetry using a cloud computing web application.

    PubMed

    Calvo-Ortega, Juan-Francisco; Pozo, Miquel; Moragues, Sandra; Casals, Joan

    2017-07-01

    To investigate the feasibility of a fast protocol for radiochromic film dosimetry to verify intensity-modulated radiotherapy (IMRT) plans. EBT3 film dosimetry was conducted in this study using the triple-channel method implemented in the cloud computing application (Radiochromic.com). We described a fast protocol for radiochromic film dosimetry to obtain measurement results within 1h. Ten IMRT plans were delivered to evaluate the feasibility of the fast protocol. The dose distribution of the verification film was derived at 15, 30, 45min using the fast protocol and also at 24h after completing the irradiation. The four dose maps obtained per plan were compared using global and local gamma index (5%/3mm) with the calculated one by the treatment planning system. Gamma passing rates obtained for 15, 30 and 45min post-exposure were compared with those obtained after 24h. Small differences respect to the 24h protocol were found in the gamma passing rates obtained for films digitized at 15min (global: 99.6%±0.9% vs. 99.7%±0.5%; local: 96.3%±3.4% vs. 96.3%±3.8%), at 30min (global: 99.5%±0.9% vs. 99.7%±0.5%; local: 96.5%±3.2% vs. 96.3±3.8%) and at 45min (global: 99.2%±1.5% vs. 99.7%±0.5%; local: 96.1%±3.8% vs. 96.3±3.8%). The fast protocol permits dosimetric results within 1h when IMRT plans are verified, with similar results as those reported by the standard 24h protocol. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  13. On the experimental validation of model-based dose calculation algorithms for 192Ir HDR brachytherapy treatment planning

    NASA Astrophysics Data System (ADS)

    Pappas, Eleftherios P.; Zoros, Emmanouil; Moutsatsos, Argyris; Peppa, Vasiliki; Zourari, Kyveli; Karaiskos, Pantelis; Papagiannis, Panagiotis

    2017-05-01

    There is an acknowledged need for the design and implementation of physical phantoms appropriate for the experimental validation of model-based dose calculation algorithms (MBDCA) introduced recently in 192Ir brachytherapy treatment planning systems (TPS), and this work investigates whether it can be met. A PMMA phantom was prepared to accommodate material inhomogeneities (air and Teflon), four plastic brachytherapy catheters, as well as 84 LiF TLD dosimeters (MTS-100M 1  ×  1  ×  1 mm3 microcubes), two radiochromic films (Gafchromic EBT3) and a plastic 3D dosimeter (PRESAGE). An irradiation plan consisting of 53 source dwell positions was prepared on phantom CT images using a commercially available TPS and taking into account the calibration dose range of each detector. Irradiation was performed using an 192Ir high dose rate (HDR) source. Dose to medium in medium, Dmm , was calculated using the MBDCA option of the same TPS as well as Monte Carlo (MC) simulation with the MCNP code and a benchmarked methodology. Measured and calculated dose distributions were spatially registered and compared. The total standard (k  =  1) spatial uncertainties for TLD, film and PRESAGE were: 0.71, 1.58 and 2.55 mm. Corresponding percentage total dosimetric uncertainties were: 5.4-6.4, 2.5-6.4 and 4.85, owing mainly to the absorbed dose sensitivity correction and the relative energy dependence correction (position dependent) for TLD, the film sensitivity calibration (dose dependent) and the dependencies of PRESAGE sensitivity. Results imply a LiF over-response due to a relative intrinsic energy dependence between 192Ir and megavoltage calibration energies, and a dose rate dependence of PRESAGE sensitivity at low dose rates (<1 Gy min-1). Calculations were experimentally validated within uncertainties except for MBDCA results for points in the phantom periphery and dose levels  <20%. Experimental MBDCA validation is laborious, yet feasible. Further

  14. SU-G-TeP4-15: The Roucoulette: A Set of Quality Control Tests for Dynamic Trajectory (4Pi) Treatment Delivery Techniques

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Teke, T

    Purpose: To present and validate a set of quality control tests for trajectory treatment delivery using synchronized dynamic couch (translation and rotation), MLC and collimator motion. Methods: The quality control tests are based on the Picket fence test, which consist of 5 narrow band 2mm width spaced at 2.5cm intervals, and adds progressively synchronized dynamic motions. The tests were exposed on GafChromic EBT3 films. The first test is a regular (no motion and MLC static while beam is on) Picket Fence test used as baseline. The second test includes simultaneous collimator and couch rotation, each stripe corresponding to a differentmore » rotation speed. Errors in these tests were introduced (0.5 degree and 1 degree error in rotation synchronization) to assess the error sensitivity of this test. The second test is similar to the regular Picket Fence but now including dynamic MLC motion and couch translation (including acceleration during delivery) while the beam is on. Finally in the third test, which is a combination of the first and second test, the Picket Fence pattern is delivered using synchronized collimator and couch rotation and synchronized dynamic MLC and couch translation including acceleration. Films were analyzed with FilmQA Pro. Results: The distance between the peaks in the dose profile where measured (18.5cm away from the isocentre in the inplane direction where non synchronized rotation would have the largest effect) and compared to the regular Picket Fence tests. For well synchronized motions distances between peaks where between 24.9–25.4 mm identical to the regular Picket Fence test. This range increased to 24.4–26.4mm and 23.4–26.4mm for 0.5 degree and 1 degree error respectively. The amplitude also decreased up to 15% when errors are introduced. Conclusion: We demonstrated that the Roucoulette tests can be used as a quality control tests for trajectory treatment delivery using synchronized dynamic motion.« less

  15. Simulation of therapeutic electron beam tracking through a non-uniform magnetic field using finite element method

    PubMed Central

    Tahmasebibirgani, Mohammad Javad; Maskani, Reza; Behrooz, Mohammad Ali; Zabihzadeh, Mansour; Shahbazian, Hojatollah; Fatahiasl, Jafar; Chegeni, Nahid

    2017-01-01

    Introduction In radiotherapy, megaelectron volt (MeV) electrons are employed for treatment of superficial cancers. Magnetic fields can be used for deflection and deformation of the electron flow. A magnetic field is composed of non-uniform permanent magnets. The primary electrons are not mono-energetic and completely parallel. Calculation of electron beam deflection requires using complex mathematical methods. In this study, a device was made to apply a magnetic field to an electron beam and the path of electrons was simulated in the magnetic field using finite element method. Methods A mini-applicator equipped with two neodymium permanent magnets was designed that enables tuning the distance between magnets. This device was placed in a standard applicator of Varian 2100 CD linear accelerator. The mini-applicator was simulated in CST Studio finite element software. Deflection angle and displacement of the electron beam was calculated after passing through the magnetic field. By determining a 2 to 5cm distance between two poles, various intensities of transverse magnetic field was created. The accelerator head was turned so that the deflected electrons became vertical to the water surface. To measure the displacement of the electron beam, EBT2 GafChromic films were employed. After being exposed, the films were scanned using HP G3010 reflection scanner and their optical density was extracted using programming in MATLAB environment. Displacement of the electron beam was compared with results of simulation after applying the magnetic field. Results Simulation results of the magnetic field showed good agreement with measured values. Maximum deflection angle for a 12 MeV beam was 32.9° and minimum deflection for 15 MeV was 12.1°. Measurement with the film showed precision of simulation in predicting the amount of displacement in the electron beam. Conclusion A magnetic mini-applicator was made and simulated using finite element method. Deflection angle and displacement

  16. SU-E-T-604: Penumbra Characteristics of a New InCiseâ„¢ Multileaf Collimator of CyberKnife M6â„¢ System

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hwang, M; Jang, S; Ozhasoglu, C

    2015-06-15

    Purpose: The InCise™ Multileaf Collimator (MLC) of CyberKnife M6™ System has been released recently. The purpose of this study was to explore the dosimetric characteristics of the new MLC. In particular, the penumbra characteristics of MLC fields at varying locations are evaluated. Methods: EBT3-based film measurements were performed with varying MLC fields ranging from 7.5 mm to 27.5 mm. Seventeen regions of interests (ROIs) were identified for irradiation. These are regions located at the central area (denoted as reference field), at the left/right edge areas of reference open field, at an intermediate location between central and edge area. Single beammore » treatment plans were designed by using the MultiPlan and was delivered using the Blue Phantom. Gafchromic films were irradiated at 1.5 cm depth in the Blue Phantom and analyzed using the Film Pro software. Variation of maximum dose, penumbra of MLC-defined fields, and symmetry/flatness were calculated as a function of locations of MLC fields. Results: The InCise™ MLC System showed relatively consistent dose distribution and penumbra size with varying locations of MLC fields. The measured maximum dose varied within 5 % at different locations compared to that at the central location and agreed with the calculated data well within 2%. The measured penumbrae were in the range of 2.9 mm and 3.7 mm and were relatively consistent regardless of locations. However, dose profiles in the out-of-field and in-field regions varied with locations and field sizes. Strong variation was seen for all fields located at 55 mm away from the central field. The MLC leakage map showed that the leakage is dependent on position. Conclusion: The size of penumbra and normalized maximum dose for MLC-defined fields were consistent in different regions of MLC. However, dose profiles in the out-field region varied with locations and field sizes.« less

  17. WE-G-BRD-08: End-To-End Targeting Accuracy of the Gamma Knife for Trigeminal Neuralgia

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Brezovich, I; Wu, X; Duan, J

    2014-06-15

    Purpose: Current QA procedures verify accuracy of individual equipment parameters, but may not include CT and MRI localizers. This study uses an end-to-end approach to measure the overall targeting errors in individual patients previously treated for trigeminal neuralgia. Methods: The trigeminal nerve is simulated by a 3 mm long, 3.175 mm (1/8 inch) diameter MRI contrast-filled cavity embedded within a PMMA plastic capsule. The capsule is positioned within the head frame such that the cavity position matches the Gamma Knife coordinates of 10 previously treated patients. Gafchromic EBT2 film is placed at the center of the cavity in coronal andmore » sagittal orientations. The films are marked with a pin prick to identify the cavity center. Treatments are planned for delivery with 4 mm collimators using MRI and CT scans acquired with the clinical localizer boxes and acquisition protocols. Coordinates of shots are chosen so that the cavity is centered within the 50% isodose volume. Following irradiation, the films are scanned and analyzed. Targeting errors are defined as the distance between the pin prick and the centroid of the 50% isodose line. Results: Averaged over 10 patient simulations, targeting errors along the x, y and z coordinates (patient left-to-right, posterior-anterior, head-to-foot) were, respectively, −0.060 +/− 0.363, −0.350 +/− 0.253, and 0.364 +/− 0.191 mm when MRI was used for treatment planning. Planning according to CT exhibited generally smaller errors, namely 0.109 +/− 0.167, −0.191 +/− 0.144, and 0.211 +/− 0.94 mm. The largest errors in MRI and CT planned treatments were, respectively, y = −0.761 and x = 0.428 mm. Conclusion: Unless patient motion or stronger MRI image distortion in actual treatments caused additional errors, all patients received the prescribed dose, i.e., the targeted section of the trig±eminal nerve was contained within the 50% isodose surface in all cases.« less

  18. SU-F-T-93: Breast Surface Dose Enhancement Using a Clinical Prone Breast Board

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Guerra, M; Jozsef, G

    Purpose: The use of specialized patient set-up devices in radiotherapy, such as prone breast boards, may have unwanted dosimetric effects. The goal of this study was to evaluate the effect of a clinically used prone breast board on skin dose due to buildup. Methods: GafChromic film (EBT3) was used for dose measurements on the surface of a solid water phantom shaped to mimic the curvature of the breast. We investigated two setup scenarios: the medial field border placed at the medial edge of the board and 1 cm contralaterally from that edge. A strip of film was taped to themore » medial surface of the phantom. Gantry angles varied from 10 to 30 degrees below the lateral gantry position, representing anterior oblique fields. The measurements were performed with and without the presence of the board; the ratio of their corresponding doses (dose enhancement) was evaluated. Results: For the cases where the field edge is at the edge of the board, the dose enhancement is negligible for all the tested angles. When the field edge is 1 cm inside the board, the maximum surface dose enhancement varies depending on the gantry angle between 2.2 for 30 degrees and 3.2 for 20 degrees. The length on the film at which the presence of the board is detectable (i.e. where there is dose enhancement) is longer for the shallower angles. Conclusion: Even the low-density, thin carbon fiber board with a thin soft foam pad on the top can produce significant dose enhancement on the skin in prone breast treatment due to loss of buildup. However, it happens only when the patient mid-sternum is over the board, i.e. the medial edge of the field traverses through the board and pad. Even then, the effect occurs only at the field edge, i.e. the penumbral region.« less

  19. Evaluation of a commercially‐available block for spatially fractionated radiation therapy

    PubMed Central

    Buckey, Courtney; Cashon, Ken; Gutierrez, Alonso; Esquivel, Carlos; Shi, Chengyu; Papanikolaou, Nikos

    2010-01-01

    In this paper, we present the dosimetric characteristics of a commercially‐produced universal GRID block for spatially fractioned radiation therapy. The dosimetric properties of the GRID block were evaluated. Ionization chamber and film measurements using both Kodak EDR2 and Gafchromic EBT film were performed in a solid water phantom to determine the relative output of the GRID block as well as its spatial dosimetric characteristics. The surface dose under the block and at the openings was measured using ultra thin TLDs. After introducing the GRID block into the treatment planning system, a treatment plan was created using the GRID block and also by creating a GRID pattern using the multi‐leaf collimator. The percent depth doses measured with film showed that there is a shift of the dmax towards shallower depths for both energies (6 MV and 18 MV) under investigation. It was observed that the skin dose at the GRID openings was higher than the corresponding open field by a factor as high as 50% for both photon energies. The profiles showed the transmission under the block was in the order of 15–20% for 6 MV and 30% for 18 MV. The MUs calculated for a real patient using the block were about 80% less than the corresponding MUs for the same plan using the multileaf collimator to define the GRID. Based on this investigation, this brass GRID compensator is a viable alternative to other solid compensators or MLC‐based fields currently in use. Its ease of creation and use give it decided advantages. Its ability to be created once and used for multiple patients (by varying the collimation of the linear accelerator jaws) makes it attractive from a cost perspective. We believe this compensator can be put to clinical use, and will allow more centers to offer GRID therapy to their patients. PACS number: 87.53.Mr

  20. Independent Monte-Carlo dose calculation for MLC based CyberKnife radiotherapy

    NASA Astrophysics Data System (ADS)

    Mackeprang, P.-H.; Vuong, D.; Volken, W.; Henzen, D.; Schmidhalter, D.; Malthaner, M.; Mueller, S.; Frei, D.; Stampanoni, M. F. M.; Dal Pra, A.; Aebersold, D. M.; Fix, M. K.; Manser, P.

    2018-01-01

    This work aims to develop, implement and validate a Monte Carlo (MC)-based independent dose calculation (IDC) framework to perform patient-specific quality assurance (QA) for multi-leaf collimator (MLC)-based CyberKnife® (Accuray Inc., Sunnyvale, CA) treatment plans. The IDC framework uses an XML-format treatment plan as exported from the treatment planning system (TPS) and DICOM format patient CT data, an MC beam model using phase spaces, CyberKnife MLC beam modifier transport using the EGS++ class library, a beam sampling and coordinate transformation engine and dose scoring using DOSXYZnrc. The framework is validated against dose profiles and depth dose curves of single beams with varying field sizes in a water tank in units of cGy/Monitor Unit and against a 2D dose distribution of a full prostate treatment plan measured with Gafchromic EBT3 (Ashland Advanced Materials, Bridgewater, NJ) film in a homogeneous water-equivalent slab phantom. The film measurement is compared to IDC results by gamma analysis using 2% (global)/2 mm criteria. Further, the dose distribution of the clinical treatment plan in the patient CT is compared to TPS calculation by gamma analysis using the same criteria. Dose profiles from IDC calculation in a homogeneous water phantom agree within 2.3% of the global max dose or 1 mm distance to agreement to measurements for all except the smallest field size. Comparing the film measurement to calculated dose, 99.9% of all voxels pass gamma analysis, comparing dose calculated by the IDC framework to TPS calculated dose for the clinical prostate plan shows 99.0% passing rate. IDC calculated dose is found to be up to 5.6% lower than dose calculated by the TPS in this case near metal fiducial markers. An MC-based modular IDC framework was successfully developed, implemented and validated against measurements and is now available to perform patient-specific QA by IDC.

  1. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Poppinga, D., E-mail: daniela.poppinga@uni-oldenburg.de; Schoenfeld, A. A.; Poppe, B.

    Purpose: The purpose of this study is the correction of the lateral scanner artifact, i.e., the effect that, on a large homogeneously exposed EBT3 film, a flatbed scanner measures different optical densities at different positions along thex axis, the axis parallel to the elongated light source. At constant dose, the measured optical densitiy profiles along this axis have a parabolic shape with significant dose dependent curvature. Therefore, the effect is shortly called the parabola effect. The objective of the algorithm developed in this study is to correct for the parabola effect. Any optical density measured at given position x ismore » transformed into the equivalent optical density c at the apex of the parabola and then converted into the corresponding dose via the calibration of c versus dose. Methods: For the present study EBT3 films and an Epson 10000XL scanner including transparency unit were used for the analysis of the parabola effect. The films were irradiated with 6 MV photons from an Elekta Synergy accelerator in a RW3 slab phantom. In order to quantify the effect, ten film pieces with doses graded from 0 to 20.9 Gy were sequentially scanned at eight positions along thex axis and at six positions along the z axis (the movement direction of the light source) both for the portrait and landscape film orientations. In order to test the effectiveness of the new correction algorithm, the dose profiles of an open square field and an IMRT plan were measured by EBT3 films and compared with ionization chamber and ionization chamber array measurement. Results: The parabola effect has been numerically studied over the whole measuring field of the Epson 10000XL scanner for doses up to 20.9 Gy and for both film orientations. The presented algorithm transforms any optical density at positionx into the equivalent optical density that would be measured at the same dose at the apex of the parabola. This correction method has been validated up to doses of 5.2 Gy all

  2. Volumetric-modulated Arc Therapy Lung Stereotactic Body Radiation Therapy Dosimetric Quality Assurance: A Comparison between Radiochromic Film and Chamber Array.

    PubMed

    Colodro, Juan Fernando Mata; Berná, Alfredo Serna; Puchades, Vicente Puchades; Amores, David Ramos; Baños, Miguel Alcaraz

    2017-01-01

    The aim of this work is to verify the use of radiochromic film in the quality assurance (QA) of volumetric-modulated arc therapy (VMAT) lung stereotactic body radiation therapy (SBRT) plans and compare the results with those obtained using an ion chamber array. QA was performed for 14 plans using a two-dimensional-array seven29 and EBT3 film. Dose values per session ranged between 7.5 Gy and 18 Gy. The multichannel method was used to obtain a dose map for film. The results obtained were compared with treatment planning system calculated profiles through gamma analysis. Passing criteria were 3%/3 mm, 2%/2 mm and 3%/1.5 mm with maximum and local dose (LD) normalization. Mean gamma passing rate (GPR) (percentage of points presenting a gamma function value of <1) was obtained and compared. Calibration curves were obtained for each color channel within the dose range 0-16 Gy. Mean GPR values for film were >98.9% for all criteria when normalizing per maximum dose. When using LD, normalization was >92.7%. GPR values for the array were lower for all criteria; this difference being statistically significant when normalizing at LD, reaching 12% for the 3%/1.5 mm criterion. Both detectors provide satisfactory results for the QA of plans for VMAT lung SBRT. The film provided greater mean GPR values, afforded greater spatial resolution and was more efficient overall.

  3. Volumetric-modulated Arc Therapy Lung Stereotactic Body Radiation Therapy Dosimetric Quality Assurance: A Comparison between Radiochromic Film and Chamber Array

    PubMed Central

    Colodro, Juan Fernando Mata; Berná, Alfredo Serna; Puchades, Vicente Puchades; Amores, David Ramos; Baños, Miguel Alcaraz

    2017-01-01

    Introduction: The aim of this work is to verify the use of radiochromic film in the quality assurance (QA) of volumetric-modulated arc therapy (VMAT) lung stereotactic body radiation therapy (SBRT) plans and compare the results with those obtained using an ion chamber array. Materials and Methods: QA was performed for 14 plans using a two-dimensional-array seven29 and EBT3 film. Dose values per session ranged between 7.5 Gy and 18 Gy. The multichannel method was used to obtain a dose map for film. Results: The results obtained were compared with treatment planning system calculated profiles through gamma analysis. Passing criteria were 3%/3 mm, 2%/2 mm and 3%/1.5 mm with maximum and local dose (LD) normalization. Mean gamma passing rate (GPR) (percentage of points presenting a gamma function value of <1) was obtained and compared. Calibration curves were obtained for each color channel within the dose range 0–16 Gy. Mean GPR values for film were >98.9% for all criteria when normalizing per maximum dose. When using LD, normalization was >92.7%. GPR values for the array were lower for all criteria; this difference being statistically significant when normalizing at LD, reaching 12% for the 3%/1.5 mm criterion. Conclusion: Both detectors provide satisfactory results for the QA of plans for VMAT lung SBRT. The film provided greater mean GPR values, afforded greater spatial resolution and was more efficient overall. PMID:28974858

  4. Density characterization of radiochromic film through source axis distance (SAD) technique in linac with slab phantom for radiotherapy applications

    NASA Astrophysics Data System (ADS)

    Hariani, Yousida; Haris, Bambang

    2017-05-01

    Characterization of radiochromic film density is accomplished through Source Axis Distance (SAD) technique in a slab phantom Linac with various depths and breadths of field. Type of the film used is gafchromic RTQA2. The dose of radiation exposure of the film may cause changes in the film density. This research aims to determine the relation between the density and the dose depth through the characteristic of curves to identify the depth of the dose and particular breadth of the field as a reference for the dose of radiotherapy patients. The result shows that the higher the dose is absorbed, the darker the film will be, yet the lower the density is obtained. The dose depth is determined by measuring the amount of dose received at various depths and breadths of field using film that is placed on the slab phantom with 6 MV linac radiation and dose of 300 cGy. The variation of the depth at 1.5 cm; 4 cm; 6 cm; 8 cm; 10 cm, the field size at 4 × 4 cm2, and the dose depth at 359.7 cGy; 315.3 cGy; 281.4 cGy; 241.2 cGy; 220.5 cGy were settled. The field size 6 × 6 cm2 takes the dose depth 354.6 cGy; 314.1 cGy; 282.6 cGy; 244.5 cGy; 224.7 cGy. The field size 8 × 8 cm2 takes the dose depth 351.6 cGy; 313 cGy; 283.8 cGy; 247.2 cGy; 228 cGy. The field size 10 × 10 cm2 takes the dose depth 348.9 cGy; 342.6 cGy; 248.4 cGy; 249.6 cGy; 231 cGy.

  5. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ravindran, P; Wui Ann, W; Lim, Y

    Purpose: In general, the linear accelerator is gated using respiratory signal obtained by way of external sensors to account for the breathing motion during radiotherapy. One of the commonly used gating devices is the Varian RPM device. Calypso system that uses electromagnetic tracking of implanted or surface transponders could also be used for gating. The aim of this study is to compare the gating efficiency of RPM device and the calypso system by phantom studies. Methods: An ArcCheck insert was used as the phantom with a Gafchromic film placed in its holder. The ArcCheck insert was placed on a Motionmore » Sim platform and moved in the longitudinal direction simulating a respiratory motion with a period of 5 seconds and amplitude of ±6mm. The Gafchromic film was exposed to a 2 × 2cm{sup 2} field, i) with the phantom static, ii) phantom moving but ungated iii) gated with gating window of 2mm and 3mm. This was repeated with Calypso system using surface transponders with the same gating window. The Gafchromic films were read with an EPSON 11000 flatbed scanner and analysed with ‘Medphysto’ software. Results: The full width at half maximum (FWHM) as measured with film at the level of the film holder was 1.65cm when the phantom was static. FWHM measured with phantom moving and without gating was 1.16 cm and penumbra was 7 mm (80–20%) on both sides. When the beam was gated with 2 mm gating window the FWHM was 1.8 cm with RPM device and 1.9 cm with Calypso. Similarly, when the beam was gated with 3 mm window, the FWHM was 1.9cm with RPM device and 2cm with Calypso. Conclusion: This work suggests that the gating efficiency of RPM device is better than that of the Calypso with surface transponder, with reference to the latency in gating.« less

  6. Sensitivity and variability of Presage dosimeter formulations in sheet form with application to SBRT and SRS QA

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Dumas, Michael, E-mail: mdumas1127@gmail.com; Rakowski, Joseph T.

    Purpose: To measure sensitivity and stability of the Presage dosimeter in sheet form for various chemical concentrations over a range of clinical photon energies and examine its use for stereotactic body radiation therapy (SBRT) and stereotactic radiosurgery (SRS) QA. Methods: Presage polymer dosimeters were formulated to investigate and optimize their sensitivity and stability. The dosimeter is composed of clear polyurethane base, leucomalachite green (LMG) reporting dye, and bromoform radical initiator in 0.9–1.0 mm thick sheets. The chemicals are mixed together for 2 min, cast in an aluminum mold, and left to cure at 60 psi for a minimum of twomore » days. Dosimeter response was characterized at energies Co-60, 6 MV, 10 MV flattening-filter free, 15 MV, 50 kVp (mean 19.2 keV), and Ir-192. The dosimeters were scanned by a Microtek Scanmaker i800 at 300 dpi, 2{sup 16} bit depth per color channel. Red component images were analyzed with ImageJ and RIT. SBRT QA was done with gamma analysis tolerances of 2% and 2 mm DTA. Results: The sensitivity of the Presage dosimeter increased with increasing concentration of bromoform. Addition of tin catalyst decreased curing time and had negligible effect on sensitivity. LMG concentration should be at least as high as the bromoform, with ideal concentration being 2% wt. Gamma Knife SRS QA measurements of relative output and profile widths were within 2% of manufacturer’s values validated at commissioning, except the 4 mm collimator relative output which was within 3%. The gamma pass rate of Presage with SBRT was 73.7%, compared to 93.1% for EBT2 Gafchromic film. Conclusions: The Presage dosimeter in sheet form was capable of detecting radiation over all tested photon energies and chemical concentrations. The best sensitivity and photostability of the dosimeter were achieved with 2.5% wt. LMG and 8.2% wt. bromoform. Scanner used should not emit any UV radiation as it will expose the dosimeter, as with the Epson 10000 XL

  7. Dosimetric characterization of a new directional low-dose rate brachytherapy source.

    PubMed

    Aima, Manik; DeWerd, Larry A; Mitch, Michael G; Hammer, Clifford G; Culberson, Wesley S

    2018-05-24

    CivaTech Oncology Inc. (Durham, NC) has developed a novel low-dose rate (LDR) brachytherapy source called the CivaSheet. TM The source is a planar array of discrete elements ("CivaDots") which are directional in nature. The CivaDot geometry and design are considerably different than conventional LDR cylindrically symmetric sources. Thus, a thorough investigation is required to ascertain the dosimetric characteristics of the source. This work investigates the repeatability and reproducibility of a primary source strength standard for the CivaDot and characterizes the CivaDot dose distribution by performing in-phantom measurements and Monte Carlo (MC) simulations. Existing dosimetric formalisms were adapted to accommodate a directional source, and other distinguishing characteristics including the presence of gold shield x-ray fluorescence were addressed in this investigation. Primary air-kerma strength (S K ) measurements of the CivaDots were performed using two free-air chambers namely, the Variable-Aperture Free-Air Chamber (VAFAC) at the University of Wisconsin Medical Radiation Research Center (UWMRRC) and the National Institute of Standards and Technology (NIST) Wide-Angle Free-Air Chamber (WAFAC). An intercomparison of the two free-air chamber measurements was performed along with a comparison of the different assumed CivaDot energy spectra and associated correction factors. Dose distribution measurements of the source were performed in a custom polymethylmethacrylate (PMMA) phantom using Gafchromic TM EBT3 film and thermoluminescent dosimeter (TLD) microcubes. Monte Carlo simulations of the source and the measurement setup were performed using MCNP6 radiation transport code. The CivaDot S K was determined using the two free-air chambers for eight sources with an agreement of better than 1.1% for all sources. The NIST measured CivaDot energy spectrum intensity peaks were within 1.8% of the MC-predicted spectrum intensity peaks. The difference in the net source

  8. SU-F-T-52: Study of Energy Dependent Effect of Dosimetry Systems Used in Therapeutic Soft X-Ray Energy Range

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Souri, S; Qian, X; Gill, G

    Purpose: To investigate energy dependent effects of different dosimetry systems which can be used as in vivo dosimetry monitoring for intraoperative radiotherapy in therapeutic soft x-ray energy range. Methods: Three dosimetry systems were evaluated in therapeutic soft x-ray energy range: optically stimulated luminescent dosimeter (OSLD) nanoDots, radiochromic EBT2 and EBT3 films. The x-ray photons were produced by a Zeiss Intrabeam 50 kV x-ray radiotherapy system. Solid water and bolus slabs with different thicknesses were used in the process of irradiation. An aluminum filter set was used to measure HVLs of X-rays. Calibration curves were made at different depth of boluses.more » Results: Half Value Layers at depths of 0, 3, 10, and 20 mm of solid water were measured to represent the energy change versus depth, yielding 0.306, 0.482, 0.865 and 0.901 respectively and indicating nearly unchanged HVL beyond 1 cm depth. The responses of each system at different depths were normalized to the response at 2 cm depth. In film dosimetry, the response is calculated as optical density (OD). The results show that there is nearly the same energy dependence for EBT2 and EBT3. At a HVL of 0.482 mm Al, the relative responses of nanoDots and EBT3 are 0.85 ± 0.04 and 0.89 ± 0.03 compared to those at 0.901 mm Al HVL, respectively, indicating no obvious difference between those two systems within the measurement uncertainty. Conclusion: It was observed that the studied dosimeter response increases about 13% from the x-ray energy of 0.48 mm Al to 0.90 mm Al. Therefore, caution should be exercised in using an appropriate calibration curve, and x-ray beam hardening effect has to be taken into account.« less

  9. A new form of the calibration curve in radiochromic dosimetry. Properties and results.

    PubMed

    Tamponi, Matteo; Bona, Rossana; Poggiu, Angela; Marini, Piergiorgio

    2016-07-01

    This work describes a new form of the calibration curve for radiochromic dosimetry that depends on one fit parameter. Some results are reported to show that the new curve performs as well as those previously used and, more importantly, significantly reduces the dependence on the lot of films, the film orientation on the scanner, and the time after exposure. The form of the response curve makes use of the net optical densities ratio against the dose and has been studied by means of the Beer-Lambert law and a simple modeling of the film. The new calibration curve has been applied to EBT3 films exposed at 6 and 15 MV energy beams of linear accelerators and read-out in transmission mode by means of a flatbed color scanner. Its performance has been compared to that of two established forms of the calibration curve, which use the optical density and the net optical density against the dose. Four series of measurements with four lots of EBT3 films were used to evaluate the precision, accuracy, and dependence on the time after exposure, orientation on the scanner and lot of films. The new calibration curve is roughly subject to the same dose uncertainty, about 2% (1 standard deviation), and has the same accuracy, about 1.5% (dose values between 50 and 450 cGy), as the other calibration curves when films of the same lot are used. Moreover, the new calibration curve, albeit obtained from only one lot of film, shows a good agreement with experimental data from all other lots of EBT3 films used, with an accuracy of about 2% and a relative dose precision of 2.4% (1 standard deviation). The agreement also holds for changes of the film orientation and of the time after exposure. The dose accuracy of this new form of the calibration curve is always equal to or better than those obtained from the two types of curves previously used. The use of the net optical densities ratio considerably reduces the dependence on the lot of films, the landscape/portrait orientation, and the time

  10. SU-E-T-188: Film Dosimetry Verification of Monte Carlo Generated Electron Treatment Plans

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Enright, S; Asprinio, A; Lu, L

    2014-06-01

    Purpose: The purpose of this study was to compare dose distributions from film measurements to Monte Carlo generated electron treatment plans. Irradiation with electrons offers the advantages of dose uniformity in the target volume and of minimizing the dose to deeper healthy tissue. Using the Monte Carlo algorithm will improve dose accuracy in regions with heterogeneities and irregular surfaces. Methods: Dose distributions from GafChromic{sup ™} EBT3 films were compared to dose distributions from the Electron Monte Carlo algorithm in the Eclipse{sup ™} radiotherapy treatment planning system. These measurements were obtained for 6MeV, 9MeV and 12MeV electrons at two depths. Allmore » phantoms studied were imported into Eclipse by CT scan. A 1 cm thick solid water template with holes for bonelike and lung-like plugs was used. Different configurations were used with the different plugs inserted into the holes. Configurations with solid-water plugs stacked on top of one another were also used to create an irregular surface. Results: The dose distributions measured from the film agreed with those from the Electron Monte Carlo treatment plan. Accuracy of Electron Monte Carlo algorithm was also compared to that of Pencil Beam. Dose distributions from Monte Carlo had much higher pass rates than distributions from Pencil Beam when compared to the film. The pass rate for Monte Carlo was in the 80%–99% range, where the pass rate for Pencil Beam was as low as 10.76%. Conclusion: The dose distribution from Monte Carlo agreed with the measured dose from the film. When compared to the Pencil Beam algorithm, pass rates for Monte Carlo were much higher. Monte Carlo should be used over Pencil Beam for regions with heterogeneities and irregular surfaces.« less

  11. Comparison of Chest Wall and Lymphatic Radiotherapy Techniques in Patients with Left Breast Carcinoma.

    PubMed

    Gültekin, Melis; Karabuğa, Mehmet; Yıldız, Ferah; Özyiğit, Gökhan; Cengiz, Mustafa; Zorlu, Faruk; Akyol, Fadıl; Gürkaynak, Murat

    2014-04-01

    The aim of this study was to find the most appropriate technique for postmastectomy chest wall (CW) and lymphatic irradiation. Partially wide tangent, 30/70 photon/electron mix, 20/80 photon/electron mix and CW and internal mammary en face electron field, were studied on computerized tomography (CT) scans of 10 left breast carcinoma patients and dosimetric calculations have been studied. Dose volume histograms (DVH) obtained from treatment planning system (TPS) were used for minimal, maximal and mean doses received by the clinical target volumes and critical structures. Partially wide tangent field resulted in the most homogeneous dose distribution for the CW and a significantly lower lung and heart doses compared with all other techniques. However, right breast dose was significantly higher for partially wide tangent technique than that each of the other techniques. Approximately 0.6-7.9% differences were found between thermoluminescent dosimeter (TLD) and treatment planning system (TPS). The daily surface doses calculating using Gafchromic® external beam therapy (EBT) dosimetry films were 161.8±2.7 cGy for the naked, 241.0±1.5 cGy when 0.5 cm bolus was used and 255.3±2.7 cGy when 1 cm bolus was used. As a result of this study, partially wide tangent field was found to be the most appropriate technique in terms of the dose distribution, treatment planning and set-up procedure. The main disadvantage of this technique was the higher dose to the contralateral breast comparing the other techniques.

  12. Sci-Sat AM(2): Brachy-05: Dosimetry effects of the TG-43 approximations for two iodine seeds in LDR brachytherapy.

    PubMed

    Furstoss, C; Bertrand, M J; Poon, E; Reniers, B; Pignol, J P; Carrier, J F; Beaulieu, L; Verhaegen, F

    2008-07-01

    This work consists of studying the interseed and tissue composition effects for two model iodine seeds: the IBt Interseed-125 and the 6711 model seed. Three seeds were modeled with the MCNP MC code in a water sphere to evaluate the interseed effect. The dose calculated at different distances from the centre was compared to the dose summed when the seeds were simulated separately. The tissue composition effect was studied calculating the radial dose function for different tissues. Before carrying out post-implant studies, the absolute dose calculated by MC was compared to experiment results: with LiF TLDs in an acrylic breast phantom and with an EBT Gafchromic film placed in a water tank. Afterwards, the TG-43 approximation effects were studied for a prostate and breast post-implant. The interseed effect study shows that this effect is more important for model 6711 (15%) than for IBt (10%) due to the silver rod in 6711. For both seed models the variations of the radial dose function as a function of the tissue composition are quasi similar. The absolute dose comparisons between MC calculations and experiments give good agreement (inferior to 3% in general). For the prostate and breast post-implant studies, a 10% difference between MC calculations and the TG-43 is found for both models of seeds. This study shows that the differences in dose distributions between TG43 and MC are quite similar for the two models of seeds and are about 10% for the studied post-implant treatments. © 2008 American Association of Physicists in Medicine.

  13. Validation of a pretreatment delivery quality assurance method for the CyberKnife Synchrony system.

    PubMed

    Mastella, E; Vigorito, S; Rondi, E; Piperno, G; Ferrari, A; Strata, E; Rozza, D; Jereczek-Fossa, B A; Cattani, F

    2016-08-01

    To evaluate the geometric and dosimetric accuracies of the CyberKnife Synchrony respiratory tracking system (RTS) and to validate a method for pretreatment patient-specific delivery quality assurance (DQA). An EasyCube phantom was mounted on the ExacTrac gating phantom, which can move along the superior-inferior (SI) axis of a patient to simulate a moving target. The authors compared dynamic and static measurements. For each case, a Gafchromic EBT3 film was positioned between two slabs of the EasyCube, while a PinPoint ionization chamber was placed in the appropriate space. There were three steps to their evaluation: (1) the field size, the penumbra, and the symmetry of six secondary collimators were measured along the two main orthogonal axes. Dynamic measurements with deliberately simulated errors were also taken. (2) The delivered dose distributions (from step 1) were compared with the planned ones, using the gamma analysis method. The local gamma passing rates were evaluated using three acceptance criteria: 3% local dose difference (LDD)/3 mm, 2%LDD/2 mm, and 3%LDD/1 mm. (3) The DQA plans for six clinical patients were irradiated in different dynamic conditions, to give a total of 19 cases. The measured and planned dose distributions were evaluated with the same gamma-index criteria used in step 2 and the measured chamber doses were compared with the planned mean doses in the sensitive volume of the chamber. (1) A very slight enlargement of the field size and of the penumbra was observed in the SI direction (on average <1 mm), in line with the overall average CyberKnife system error for tracking treatments. (2) Comparison between the planned and the correctly delivered dose distributions confirmed the dosimetric accuracy of the RTS for simple plans. The multicriteria gamma analysis was able to detect the simulated errors, proving the robustness of their method of analysis. (3) All of the DQA clinical plans passed the tests, both in static and dynamic conditions

  14. SU-G-TeP2-08: Evaluation of Plastic Scintillator Detector for Small Field Stereotactic Patient-Specific Quality Assurance

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Qin, Y; Gardner, S; Huang, Y

    Purpose: To evaluate the performance of a commercial plastic scintillator detector (PSD) for small-field stereotactic patient-specific quality assurance using flattening-filter-free (FFF) beams. Methods: A total of ten spherical targets (volume range:[0.03cc–2cc]) were planned using Dynamic Conformal Arc(DCA-10 plans) and Volumetric Modulated Arc Therapy(VMAT-10 plans) techniques in Eclipse(AAA v.11, 1mm dose calculation grid size). Additionally, 15 previously-treated cranial and spine SRS plans were evaluated (6 DCA, 9 VMAT, volume range:[0.04cc–119.02cc]). All measurements were acquired using Varian Edge equipped with HDMLC. Three detectors were used: PinPoint ion chamber (PTW;active volume 0.015cc), Exradin W1 PSD (Standard Imaging;active volume 0.002cc), and Gafchromic EBT3 filmmore » (Ashland). PinPoint and PSD were positioned perpendicular to beam axis in a Lucy phantom (Standard Imaging). Films were placed at isocenter in solid water. Calibration films were delivered for absolute dose analysis. Results: For large spherical targets(>1.5cc) with DCA, all detectors agreed within 1% of AAA calculations. As target volume decreased, PSD measured higher doses than AAA (maximum difference: 3.3% at 0.03cc target), while PinPoint chamber measured lower doses (maximum difference:-3.8% at 0.03cc target). Inter-detector differences between pinpoint and PSD increased with decreasing target size; differences>5% were observed for targets<0.09cc. Similar trends for inter-detector behavior were observed for clinical plans. For target sizes<0.08cc, PSD measured>5% higher dose than PinPoint chamber (maximum difference: 9.25% at 0.04cc target). Film demonstrated agreement of −0.19±1.47% with PSD for all spherical targets, and agreement within −0.98±2.25% for all 15 clinical targets. Unlike DCA, VMAT plans did not show improved AAA-to-detector agreements for large targets. Conclusion: For all targets, the PSD measurements agreed with film within 1.0%, on average. For

  15. A new form of the calibration curve in radiochromic dosimetry. Properties and results

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Tamponi, Matteo, E-mail: mtamponi@aslsassari.it; B

    Purpose: This work describes a new form of the calibration curve for radiochromic dosimetry that depends on one fit parameter. Some results are reported to show that the new curve performs as well as those previously used and, more importantly, significantly reduces the dependence on the lot of films, the film orientation on the scanner, and the time after exposure. Methods: The form of the response curve makes use of the net optical densities ratio against the dose and has been studied by means of the Beer–Lambert law and a simple modeling of the film. The new calibration curve hasmore » been applied to EBT3 films exposed at 6 and 15 MV energy beams of linear accelerators and read-out in transmission mode by means of a flatbed color scanner. Its performance has been compared to that of two established forms of the calibration curve, which use the optical density and the net optical density against the dose. Four series of measurements with four lots of EBT3 films were used to evaluate the precision, accuracy, and dependence on the time after exposure, orientation on the scanner and lot of films. Results: The new calibration curve is roughly subject to the same dose uncertainty, about 2% (1 standard deviation), and has the same accuracy, about 1.5% (dose values between 50 and 450 cGy), as the other calibration curves when films of the same lot are used. Moreover, the new calibration curve, albeit obtained from only one lot of film, shows a good agreement with experimental data from all other lots of EBT3 films used, with an accuracy of about 2% and a relative dose precision of 2.4% (1 standard deviation). The agreement also holds for changes of the film orientation and of the time after exposure. Conclusions: The dose accuracy of this new form of the calibration curve is always equal to or better than those obtained from the two types of curves previously used. The use of the net optical densities ratio considerably reduces the dependence on the lot of films

  16. SU-F-T-669: Commissioning of An Electronic Brachytherapy System for Targeted Mouse Irradiation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Culberson, W; Micka, J; Carchman, E

    Purpose: The aim of this study was to commission the Xoft Axxent™ electronic brachytherapy (eBT) source and 10 mm diameter surface applicator with NIST traceability for targeted irradiations of mouse anal carcinomas. Methods: The Xoft Axxent™ electronic brachytherapy (eBT) and 10 mm diameter surface applicator was chosen by the collaborating physician as a radiation delivery mechanism for mouse anal carcinomas. The target dose was 2 Gy at a depth of 3 mm in tissue to be delivered in a single fraction. To implement an accurate and reliable irradiation plan, the system was commissioned by first determining the eBT source outputmore » and corresponding dose rate at a depth of 3 mm in tissue. This was determined through parallel-plate ion chamber measurements and published conversion factors. Well-type ionization chamber measurements were used to determine a transfer coefficient, which correlates the measured dose rate at 3 mm to the NIST-traceable quantity, air-kerma rate at 50 cm in air, for eBT sources. By correlating these two quantities, daily monitoring in the well chamber becomes an accurate and efficient quality assurance technique. Once the dose-rate was determined, a treatment recipe was developed and confirmed with chamber measurements to deliver the requested dose. Radiochromic film was used to verify the dose distribution across the field. Results: Dose rates at 3 mm depth in tissue were determined for two different Xoft Axxent™ sources and correlated with NIST-traceable well-type ionization chamber measurements. Unique transfer coefficients were determined for each source and the treatment recipe was validated by measurements. Film profiles showed a uniform dose distribution across the field. Conclusion: A Xoft Axxent™ eBT system was successfully commissioned for use in the irradiation of mouse rectal tumors. Dose rates in tissue were determined as well as other pertinent parameters to ensure accurate delivery of dose to the target region.« less

  17. SU-F-T-323: A Post-Mastectomy Radiation Therapy Dose Distribution Study Using Nanodots and Films

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Qian, X; Vaidya, K; Puckett, L

    Purpose: In post-mastectomy radiation therapy (RT), skin dose must be accurately estimated to assess skin reactions such as erythema, desquamation and necrosis. Planning systems cannot always provide accurate dosimetry for target volumes distal to skin. Therefore, in-vivo dosimetry is necessary. A female anthropomorphic phantom was used with optically stimulated luminescence dosimeters (nanoDots) to measure dose to chest wall skin. In addition, EBT2 films was employed to measure dose to left lung and heart in post-mastectomy RT. Methods: Films and nanoDots were calibrated under full buildup conditions at 100cm SAD for 6MV photons. Five pieces of films were placed between slabsmore » of Rando phantom to assess dose to left lung and heart. Two layers of 0.5cm thick bolus were used to cover the whole left chest. Six pairs of nanoDots were placed at medical and lateral aspects on the bolus surface, between the 0.5cm bolus layers, and under the bolus. Three control nanoDots were placed on chest wall to quantify imaging dose. The phantom was CT scanned with all dosimeters in place, and treatment planning was performed with tangential fields (200cGy). All dosimeters were contoured on CT and dose was extracted. NanoDots were read using nanoDot reader and films were scanned using film scanner. The measured and calculated doses were tabulated. Results: Dose to 12 nanoDots were evaluated. Dose variance for surface nanoDots were +3.8%, +2.7%, −5% and −9.8%. Those at lateral positions, with greater beam obliquity had larger variance than the medial positions. A similar trend was observed for other nanoDots (Table1). Point doses from films for heart and the left lung were 112.7cGy and 108.7cGy, with +10.2% and +9.04% deviation from calculated values, respectively. Conclusion: Dosimetry provided by the advanced planning system was verified using NanoDots and films. Both nanoDots and films provided good estimation of dose distribution in post-mastectomy RT.« less

  18. IMRT and RapidArc commissioning of a TrueBeam linear accelerator using TG-119 protocol cases.

    PubMed

    Wen, Ning; Zhao, Bo; Kim, Jinkoo; Chin-Snyder, Karen; Bellon, Maria; Glide-Hurst, Carri; Barton, Kenneth; Chen, Daiquan; Chetty, Indrin J

    2014-09-08

    The purpose of this study is to evaluate the overall accuracy of intensity-modulated radiation therapy (IMRT) and RapidArc delivery using both flattening filter (FF) and flattening filter-free (FFF) modalities based on test cases developed by AAPM Task Group 119. Institutional confidence limits (CLs) were established as the baseline for patient specific treatment plan quality assurance (QA). The effects of gantry range, gantry speed, leaf speed, dose rate, as well as the capability to capture intentional errors, were evaluated by measuring a series of Picket Fence (PF) tests using the electronic portal imaging device (EPID) and EBT3 films. Both IMRT and RapidArc plans were created in a Solid Water phantom (30 × 30 × 15 cm3) for the TG-119 test cases representative of normal clinical treatment sites for all five photon energies (6X, 10X, 15X, 6X-FFF, 10X-FFF) and the Exact IGRT couch was included in the dose calculation. One high-dose point in the PTV and one low-dose point in the avoidance structure were measured with an ion chamber in each case for each energy. Similarly, two GAFCHROMIC EBT3 films were placed in the coronal planes to measure planar dose distributions in both high- and low-dose regions. The confidence limit was set to have 95% of the measured data fall within the tolerance. The mean of the absolute dose deviation for variable dose rate and gantry speed during RapidArc delivery was within 0.5% for all energies. The corresponding results for leaf speed tests were all within 0.4%. The combinations of dynamic leaf gap (DLG) and MLC transmission factor were optimized based on the ion chamber measurement results of RapidArc delivery for each energy. The average 95% CLs for the high-dose point in the PTV were 0.030 ± 0.007 (range, 0.022-0.038) for the IMRT plans and 0.029 ± 0.011 (range, 0.016-0.043) for the RapidArc plans. For low-point dose in the avoidance structures, the CLs were 0.029 ± 0.006 (range, 0.024-0.039) for the IMRT plans and 0.027

  19. Surface dose measurement for helical tomotherapy.

    PubMed

    Snir, Jonatan A; Mosalaei, Homeira; Jordan, Kevin; Yartsev, Slav

    2011-06-01

    To compare the surface dose measurements made by different dosimeters for the helical tomotherapy (HT) plan in the case of the target close to the surface. Surface dose measurements in different points for the HT plan to deliver 2 Gy to the planning target volume (PTV) at 5 mm below the surface of the cylindrical phantom were performed by radiochromic films, single use metal oxide semiconductor field-effect transistor (MOSFET) dosimeters, silicon IVD QED diode, and optically stimulated luminescence (OSL) dosimeters. The measured doses by all dosimeters were within 12 +/- 8% difference of each other. Radiochromic films, EBT, and EBT2, provide high spatial resolution, although it is difficult to get accurate measurements of dose. Both the OSL and QED measured similar dose to that of the MOSFET detectors. The QED dosimeter is promising as a reusable on-line wireless dosimeter, while the OSL dosimeters are easier to use, require minimum setup time and are very precise.

  20. TU-FG-BRB-01: Dual Energy CT Proton Stopping Power Ratio Calibration and Validation with Animal Tissues

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Xie, Y; Yin, L; Ainsley, C

    Purpose: The conversion of Hounsfield Unit (HU) to proton stopping power ratio (SPR) is a main source of uncertainty in proton therapy. In this study, the SPRs of animal tissues were measured and compared with prediction from dual energy CT (DECT) and single energy CT (SECT) calibrations. Methods: A stoichiometric calibration method for DECT was applied to predict the SPR using CT images acquired at 80 kVp and 140 kVp. The dual energy index was derived based on the HUs of the paired spectral images and used to calculate the SPRs of the materials. Tissue surrogates with known chemical compositionsmore » were used for calibration, and animal tissues (pig brain, liver, kidney; veal shank, muscle) were used for validation. The materials were irradiated with proton pencil beams, and SPRs were deduced from the residual proton range measured using a multi-layer ion chamber device. In addition, Gafchromic EBT3 films were used to measure the distal dose profiles after irradiation through the tissue samples and compared with those calculated by the treatment planning system using both DECT and SECT predicted SPRs. Results: The differences in SPR between DECT prediction and measurement were −0.31±0.36% for bone, 0.47±0.42% for brain, 0.67±0.15% for liver, 0.51±0.52% for kidney, and −0.96±0.15% for muscle. The corresponding results using SECT were 3.1±0.12%, 1.90±0.45%, −0.66±0.11%, 2.33±0.21%, and −1.70±0.17%. In the film measurements, average distances between film and calculated distal dose profiles were 0.35±0.12 mm for DECT calibration and −1.22±0.12 mm for SECT calibration for a beam with a range of 15.79 cm. Conclusion: Our study indicates that DECT is superior to SECT for proton SPR prediction and has the potential to reduce the range uncertainty to less than 2%. DECT may permit the use of tighter distal and proximal range uncertainty margins for treatment, thereby increasing the precision of proton therapy.« less

  1. Penalization of aperture complexity in inversely planned volumetric modulated arc therapy

    PubMed Central

    Younge, Kelly C.; Matuszak, Martha M.; Moran, Jean M.; McShan, Daniel L.; Fraass, Benedick A.; Roberts, Donald A.

    2012-01-01

    Purpose: Apertures obtained during volumetric modulated arc therapy (VMAT) planning can be small and irregular, resulting in dosimetric inaccuracies during delivery. Our purpose is to develop and integrate an aperture-regularization objective function into the optimization process for VMAT, and to quantify the impact of using this objective function on dose delivery accuracy and optimized dose distributions. Methods: An aperture-based metric (“edge penalty”) was developed that penalizes complex aperture shapes based on the ratio of MLC side edge length and aperture area. To assess the utility of the metric, VMAT plans were created for example paraspinal, brain, and liver SBRT cases with and without incorporating the edge penalty in the cost function. To investigate the dose calculation accuracy, Gafchromic EBT2 film was used to measure the 15 highest weighted apertures individually and as a composite from each of two paraspinal plans: one with and one without the edge penalty applied. Films were analyzed using a triple-channel nonuniformity correction and measurements were compared directly to calculations. Results: Apertures generated with the edge penalty were larger, more regularly shaped and required up to 30% fewer monitor units than those created without the edge penalty. Dose volume histogram analysis showed that the changes in doses to targets, organs at risk, and normal tissues were negligible. Edge penalty apertures that were measured with film for the paraspinal plan showed a notable decrease in the number of pixels disagreeing with calculation by more than 10%. For a 5% dose passing criterion, the number of pixels passing in the composite dose distributions for the non-edge penalty and edge penalty plans were 52% and 96%, respectively. Employing gamma with 3% dose/1 mm distance criteria resulted in a 79.5% (without penalty)/95.4% (with penalty) pass rate for the two plans. Gradient compensation of 3%/1 mm resulted in 83.3%/96.2% pass rates

  2. SU-E-T-330: Dosimetric Impact of Intrafraction Respiratory Motion On Lung SBRT Treatment Using Cyberknife 0-View Tracking Mode

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rao, M; Chen, F; Cotrutz, C

    2015-06-15

    Purpose: To investigate the influence of respiratory motion on the delivered dose in lung stereotactic body radiotherapy (SBRT) using Cyberknife (CK) 0-View tracking mode. Methods: CT scans at inspiration and expiration of an anthropomorphic motion phantom were fused base on the spine and an internal target volume (ITV) was created. A 5mm expansion around the ITV resulted in the planning target volume. Three CK plans were generated in Accuray MultiPlan using Lung Optimization Tracking 0-View technique with the minimum MU per beam set to (a) 5MU, (b) 15MU and (c) 30MU, respectively. Doses were calculated on the expiration CT usingmore » Monte-Carlo algorithm. Each plan was delivered 5 times with a range of different starting phases in the respiratory cycle to assess the dose variation due to interplay effect. The delivered dose was measured with EBT3 Gafchromic film which was inserted in the moving target of the phantom. The target motion range is 3 cm in superior-inferior (SI) direction with the breathing period of 5 seconds. Results: The gamma analysis (5%/2mm) of the dose with the films in the transverse plane resulted in average passing rate of 95.5±4.1%, 96.7±2.6%, and 96.2±2.5% for plan (a), (b), and (c), respectively. For the sagittal films, the average passing rate was 91.1±4.9%, 92.1±3.6%, and 92.3±2.9% for the three plans, respectively. The disagreement between measurement and dose calculations were mostly on the target edges in SI direction. The mean measured versus calculated dose differences at the edge of target in SI direction were (a) 3.9±4.8%, (b) 2.4±3.3%, and (c) 2.2±3.2% for the three plans, respectively. Conclusions: The plans with low-MU beams (below 10MU) tend to cause slightly larger dose variation. However in terms of target coverage, the overall clinical dosimetric impact of the intrafraction respiratory motion in lung SBRT is insignificant when averaged over 3∼5 fractions.« less

  3. Development of a PET cyclotron based irradiation setup for proton radiobiology

    NASA Astrophysics Data System (ADS)

    Ghithan, Sharif; Crespo, Paulo; do Carmo, S. J. C.; Ferreira Marques, Rui; Fraga, F. A. F.; Simões, Hugo; Alves, Francisco; Rachinhas, P. J. B. M.

    2015-02-01

    An out-of-yoke irradiation setup using the proton beam from a cyclotron that ordinary produces radioisotopes for positron emission tomography (PET) has been developed, characterized, calibrated and validated. The current from a 20 μm thick aluminum transmission foil is readout by home-made transimpedance electronics, providing online dose information. The main monitoring variables, delivered in real-time, include beam current, integrated charge and dose rate. Hence the dose and integrated current delivered at a given instant to an experimental setup can be computer-controlled with a shutter. In this work, we report on experimental results and Geant4 simulations of a setup which exploits for the first time the 18 MeV proton beam from a PET cyclotron to irradiate a selected region of a target using the developed irradiation system. By using this system, we are able to deliver a homogeneous beam on targets with 18 mm diameter, allowing to achieve the controlled irradiation of cell cultures located in biological multi-well dishes of 16 mm diameter. We found that the magnetic field applied inside the cyclotron plays a major role for achieving the referred to homogeneity. The quasi-Gaussian curve obtained by scanning the magnet current and measuring the corresponding dose rate must be measured before any irradiation procedure, with the shutter closed. At the optimum magnet current, which corresponds to the center of the Gaussian, a homogenous dose is observed over the whole target area. Making use of a rotating disk with a slit of 0.5 mm at a radius of 150 mm, we could measure dose rates on target ranging from 500 mGy/s down to 5 mGy/s. For validating the developed irradiation setup, several Gafchromic® EBT2 films were exposed to different values of dose. The absolute dose in the irradiated films were assessed in the 2D film dosimetry system of the Department of Radiotherapy of Coimbra University Hospital Center with a precision better than 2%. In the future, we plan

  4. SU-F-T-86: Electron Dosimetric Effects of Bolus and Lens Shielding in Treating Superficial Eye Lesions

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Young, L; Wootton, L; Gopan, O

    Purpose: Electron therapy for the treatment of ocular lymphomas requires the lens to be shielded to prevent secondary cataracts. This work evaluates the dosimetry under a suspended eyeshield with and without bolus for low energy electron fields. Methods: Film (GafChromic EBT3) dosimetry and relative output factors were measured for 6, 8, and 10 MeV electron energies. A customized 5 cm diameter circle electron orbital cutout was constructed for a 6×6 cm applicator with a lens shield, 1 cm diameter Cerrobend cylinder with 2.2 cm length, suspended from an XV film covering the open field. Relative output factors were measured usingmore » a Scanditronix electron diode in a solid water phantom. Depth dose profiles were collected for bolus thicknesses of 0, 3, and 5 mm in solid water at a source to surface distance (SSD) of 100 cm. These measurements were repeated in a Rando phantom. Results: At 5 mm, the approximate distance of the lens from the surface of the cornea, the estimated dose in solid water under the suspended lens shield was reduced to 16%, 14%, and 13% of the unblocked dose at the same depth, for electron energies of 6, 8, and 10 MeV, respectively. Applying bolus increased estimated doses under the block to 22% for 3-mm and 32% for 5-mm thicknesses for a 6 MeV incident electron beam. This effect is reduced for higher energies where the corresponding values were 15.5% and 18% for 3-mm and 5-mm for an 8 MeV electron beam. Conclusion: The application of bolus to treat superficial eye lesions of the conjunctiva increases lens dose at a depth of 5-mm under the shielding block with decreasing electron energy. Careful selection of electron energy is needed to account for electron scatter under the lens shield with the application of bolus in order to prevent cataracts.« less

  5. Dose perturbation effect of metallic spinal implants in proton beam therapy.

    PubMed

    Jia, Yingcui; Zhao, Li; Cheng, Chee-Wai; McDonald, Mark W; Das, Indra J

    2015-09-08

    The purpose of this study was to investigate the effect of dose perturbations for two metallic spinal screw implants in proton beam therapy in the perpendicular and parallel beam geometry. A 5.5 mm (diameter) by 45 mm (length) stainless steel (SS) screw and a 5.5 mm by 35 mm titanium (Ti) screw commonly used for spinal fixation were CT-scanned in a hybrid phantom of water and solid water. The CT data were processed with an orthopedic metal artifact reduction (O-MAR) algorithm. Treatment plans were generated for each metal screw with a proton beam oriented, first parallel and then perpendicular, to the longitudinal axis of the screw. The calculated dose profiles were compared with measured results from a plane-parallel ion chamber and Gafchromic EBT2 films. For the perpendicular setup, the measured dose immediately downstream from the screw exhibited dose enhancement up to 12% for SS and 8% for Ti, respectively, but such dose perturbation was not observed outside the lateral edges of the screws. The TPS showed 5% and 2% dose reductions immediately at the interface for the SS nd Ti screws, respectively, and up to 9% dose enhancements within 1 cm outside of the lateral edges of the screws. The measured dose enhancement was only observed within 5 mm from the interface along the beam path. At deeper depths, the lateral dose profiles appeared to be similar between the measurement and TPS, with dose reduction in the screw shadow region and dose enhancement within 1-2 cm outside of the lateral edges of the metals. For the parallel setup, no significant dose perturbation was detected at lateral distance beyond 3 mm away from both screws. Significant dose discrepancies exist between TPS calculations and ion chamber and film measurements in close proximity of high-Z inhomogeneities. The observed dose enhancement effect with proton therapy is not correctly modeled by TPS. An extra measure of caution should be taken when evaluating dosimetry with spinal metallic implants.

  6. SU-F-T-490: Separating Effects Influencing Detector Response in Small MV Photon Fields

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wegener, S; Sauer, O

    2016-06-15

    Purpose: Different detector properties influence their responses especially in field sizes below the lateral electron range. Due to the finite active volume, the detector density and electron perturbation at other structural parts, the response factor is in general field size dependent. We aimed to visualize and separate the main effects contributing to detector behavior for a variety of detector types. This was achieved in an experimental setup, shielding the field center. Thus, effects caused by scattered radiation could be examined separately. Methods: Signal ratios for field sizes down to 8 mm (SSD 90 cm, water depth 10 cm) of amore » 6MV beam from a Siemens Primus LINAC were recorded with several detectors: PTW microDiamond and PinPoint ionization chamber, shielded diodes (PTW P-60008, IBA PFD and SNC Edge) and unshielded diodes (PTW E-60012 and IBA SFD). Measurements were carried out in open fields and with an aluminum pole of 4 mm diameter as a central block. The geometric volume effect was calculated from profiles obtained with Gafchromic EBT3 film, evaluated using FilmQA Pro software (Ashland, USA). Results: Volume corrections were 1.7% at maximum. After correction, in small open fields, unshielded diodes showed a lower response than the diamond, i.e. diamond detector over-response seems to be higher than that for unshielded diodes. Beneath the block, this behavior was amplified by a factor of 2. For the shielded diodes, the overresponse for small open fields could be confirmed. However their lateral response behavior was strongly type dependent, e.g. the signal ratio dropped from 1.02 to 0.98 for the P-60008 diode. Conclusion: The lateral detector response was experimentally examined. Detector volume and density alone do not fully account for the field size dependence of detector response. Detector construction details play a major role, especially for shielded diodes.« less

  7. SU-F-BRB-15: Dosimetric Study of Radiation Therapy for Head/Neck Patients with Metallic Dental Fixtures

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lu, L; Allan, E; Putten, M Van

    Purpose: To investigate the dose contributions of scattered electrons from dental amalgams during head and neck radiotherapy, and to evaluate the protective role of dosimetric dental stents during treatment to prevent oral mucositis. Methods: A phantom was produced to accurately simulate the oral cavity and head. The oral cavity consisted of a tissue equivalent upper and lower jaw and complete set of teeth. A set of 4 mm ethylene copolymer dosimetric stents was made for the upper and lower teeth. Five removable gold caps were fitted to apposing right molars, and the phantom was crafted to accomodate horizontal and verticalmore » film for 2D dosimetry and NanoDot dosimeter for recording point doses. The head was simulated using a small cylindrical glass water bath. CT simulation was performed on the phantom with and without metal fittings and, in each case, with and without the dental stent. The CT image sets were imported into Eclipse treatment planning system for contouring and treatment planning, and a 9-field IMRT treatment plan was developed for each scenario. These plans were delivered using a Varian TrueBeam linear accelerator. Doses were recorded using GafChromic EBT2 films and NanoDot dosimeters. Results: The measurements revealed a 43% relative increase in dose measured adjacent to the metal fixtures in the horizontal plane without the use of the dental stent. This equates to a total dose of 100 Gy to the oral mucosa during a standard course of definitive radiotherapy. To our knowledge, this is the first dosimetric analysis of dental stents using an anatomically realistic phantom and modern beam arrangement. Conclusion: These results support the use of dosimetric dental stents in head and neck radiotherapy for patients with metallic dental fixtures as a way to effectively reduce dose to nearby mucosal surfaces and, hence, reduce the risk and severity of mucositis.« less

  8. SU-E-T-469: A Practical Approach for the Determination of Small Field Output Factors Using Published Monte Carlo Derived Correction Factors

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Calderon, E; Siergiej, D

    2014-06-01

    Purpose: Output factor determination for small fields (less than 20 mm) presents significant challenges due to ion chamber volume averaging and diode over-response. Measured output factor values between detectors are known to have large deviations as field sizes are decreased. No set standard to resolve this difference in measurement exists. We observed differences between measured output factors of up to 14% using two different detectors. Published Monte Carlo derived correction factors were used to address this challenge and decrease the output factor deviation between detectors. Methods: Output factors for Elekta's linac-based stereotactic cone system were measured using the EDGE detectormore » (Sun Nuclear) and the A16 ion chamber (Standard Imaging). Measurements conditions were 100 cm SSD (source to surface distance) and 1.5 cm depth. Output factors were first normalized to a 10.4 cm × 10.4 cm field size using a daisy-chaining technique to minimize the dependence of field size on detector response. An equation expressing the relation between published Monte Carlo correction factors as a function of field size for each detector was derived. The measured output factors were then multiplied by the calculated correction factors. EBT3 gafchromic film dosimetry was used to independently validate the corrected output factors. Results: Without correction, the deviation in output factors between the EDGE and A16 detectors ranged from 1.3 to 14.8%, depending on cone size. After applying the calculated correction factors, this deviation fell to 0 to 3.4%. Output factors determined with film agree within 3.5% of the corrected output factors. Conclusion: We present a practical approach to applying published Monte Carlo derived correction factors to measured small field output factors for the EDGE and A16 detectors. Using this method, we were able to decrease the percent deviation between both detectors from 14.8% to 3.4% agreement.« less

  9. SU-E-T-312: Dosimetric Consideration for the Agility MLC When Planning Rotational SRT/SBRT Treatments

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kong, X; Harris, J; Spitznagel, D

    2015-06-15

    Purpose: To analyze the radiation transmission of the Agility MLC and make recommendation based on the MLC dosimetric characteristics for SRT, SBRT and VMAT planning Method and Materials: Agility MLC is the newest model from Elekta and has no back up diaphragm behind leaves for this generation. Leaves are single focused with rounded end; composed of leafs each 0.5cm wide, 9cm thick, constructed from tungsten alloy and provide low transmission <0.5%. Total radiation transmission from leaves and diaphragm is <0.13%. A 360degree arc was generated using iCom tools; leaves were programmed closed while keeping the diaphragm fully open to maximizemore » the MLC transmission effect. Gafchromic EBT films were sandwiched between 4cm of solid water and situated at midplane to take dose measurement. 5000MU was delivered using 6MV VersaHD, various collimator angles, and a 5cm central axis offset was tested also. Films were scanned with Epson 10000XL scanner and analyzed using DoseLab Pro. Results: Due to the rounded leaf end and nature of rotation therapy, dose accumulation through the leaf gap is significant. By offsetting the leaf gap from central axis, this accumulation can be greatly reduced. There are dark bands showing accumulation of interleaf transmission which is improved by increasing collimator angle from 0 to 45dgree. However for 45 degree, in most cases, there are larger volumes sweeping under MLC alone, which needs considered planning. Conclusions: While inter-leaf leakage is minimized by using collimator angles greater than 0 degrees, the location of the leaf gap must also be managed. The leaf gap position becomes critically important when the treatment area is off axis such is the case when more than one PTV is being treated. With VMAT for SRT, SBRT becoming a more popular planning technique, special attention needs to be paid when initially setting up the field geometry.« less

  10. SU-C-16A-04: Dosimetric Validation of a Partially-Shielded Gd-153 Brachytherapy Concept

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Li, X; Adams, Q; Flynn, R

    Purpose: To demonstrate by measurement that using partially shielded Gd-153 sources for rotating-shield brachytherapy (RSBT) is feasible. RSBT is a potentially superior alternative to conventional high-dose-rate brachytherapy and provides the opportunity to dramatically improve tumor dose conformity for the treatment of, for example, prostate cancer. Methods: A custom-built, stainless steel encapsulated 150 mCi Gd-153 capsule with an outer length of 12.8 mm, outer diameter of 2.10 mm, active length of 9.98 mm, and active diameter of 1.53 mm was used. A partially shielded catheter was constructed with a 500 μm platinum shield and a 500 μm aluminum emission window, bothmore » with 180° azimuthal coverage. An acrylic phantom was constructed to measure the dose distributions from the shielded catheter in the transverse plane using Gafchromic EBT3 films. Film calibration curves were generated from 50, 70, and 100 kVp x-ray beams with NIST-traceable air kerma values to account for energy variation. Results: The transmission ratios of platinum to aluminum shielding at 1 cm off-axis are 7.5% and 7.6% for Monte Carlo (MCNP5) predicted and experimental results, respectively. The predicted/measured relative dose rates at 1 cm, 2 cm and 3 cm off-axis through the Al window were 100%/92.9%, 28.6%/27.0% and 13.8%/12.7%, respectively. Through the Pt shield, the predicted/measured relative dose rates were 7.5%/7.1%, 3.8%/3.0% and 2.4%/1.7%, respectively. Conclusion: Using partially-shielded Gd-153 sources for RSBT is a promising approach to improving brachytherapy dose distributions. The next step in making Gd-153 based RSBT a reality is developing a Gd-153 source that is small enough such that the source, shield, and catheter all fit within a 16 gauge needle, which has a 1.65 mm diameter. University of Iowa Research Foundation.« less

  11. Evaluation of backscatter dose from internal lead shielding in clinical electron beams using EGSnrc Monte Carlo simulations.

    PubMed

    De Vries, Rowen J; Marsh, Steven

    2015-11-08

    Internal lead shielding is utilized during superficial electron beam treatments of the head and neck, such as lip carcinoma. Methods for predicting backscattered dose include the use of empirical equations or performing physical measurements. The accuracy of these empirical equations required verification for the local electron beams. In this study, a Monte Carlo model of a Siemens Artiste linac was developed for 6, 9, 12, and 15 MeV electron beams using the EGSnrc MC package. The model was verified against physical measurements to an accuracy of better than 2% and 2mm. Multiple MC simulations of lead interfaces at different depths, corresponding to mean electron energies in the range of 0.2-14 MeV at the interfaces, were performed to calculate electron backscatter values. The simulated electron backscatter was compared with current empirical equations to ascertain their accuracy. The major finding was that the current set of backscatter equations does not accurately predict electron backscatter, particularly in the lower energies region. A new equation was derived which enables estimation of electron backscatter factor at any depth upstream from the interface for the local treatment machines. The derived equation agreed to within 1.5% of the MC simulated electron backscatter at the lead interface and upstream positions. Verification of the equation was performed by comparing to measurements of the electron backscatter factor using Gafchromic EBT2 film. These results show a mean value of 0.997 ± 0.022 to 1σ of the predicted values of electron backscatter. The new empirical equation presented can accurately estimate electron backscatter factor from lead shielding in the range of 0.2 to 14 MeV for the local linacs.

  12. The influence of physical wedges on penumbra and in-field dose uniformity in ocular proton beams.

    PubMed

    Baker, Colin; Kacperek, Andrzej

    2016-04-01

    A physical wedge may be partially introduced into a proton beam when treating ocular tumours in order to improve dose conformity to the distal border of the tumour and spare the optic nerve. Two unwanted effects of this are observed: a predictable broadening of the beam penumbra on the wedged side of the field and, less predictably, an increase in dose within the field along a relatively narrow volume beneath the edge (toe) of the wedge, as a result of small-angle proton scatter. Monte Carlo simulations using MCNPX and direct measurements with radiochromic (GAFCHROMIC(®) EBT2) film were performed to quantify these effects for aluminium wedges in a 60 MeV proton beam as a function of wedge angle and position of the wedge relative to the patient. For extreme wedge angles (60° in eye tissue) and large wedge-to-patient distances (70 mm in this context), the 90-10% beam penumbra increased from 1.9 mm to 9.1 mm. In-field dose increases from small-angle proton scatter were found to contribute up to 21% additional dose, persisting along almost the full depth of the spread-out-Bragg peak. Profile broadening and in-field dose enhancement are both minimised by placing the wedge as close as possible to the patient. Use of lower atomic number wedge materials such as PMMA reduce the magnitude of both effects as a result of a reduced mean scattering angle per unit energy loss; however, their larger physical size and greater variation in density are undesirable. Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  13. Poster – 41: External marker block placement on the breast or chest wall for left-sided deep inspiration breath-hold radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Conroy, Leigh; Guebert, Alexandra; Smith, Wendy

    Purpose: We investigate DIBH breast radiotherapy using the Real-time Position Management (RPM) system with the marker-block placed on the target breast or chest wall. Methods: We measured surface dose for three different RPM marker-blocks using EBT3 Gafchromic film at 0° and 30° incidence. A registration study was performed to determine the breast surface position that best correlates with overall internal chest wall position. Surface and chest wall contours from MV images of the medial tangent field were extracted for 15 patients. Surface contours were divided into three potential marker-block positions on the breast: Superior, Middle, and Inferior. Translational registration wasmore » used to align the partial contours to the first-fraction contour. Each resultant transformation matrix was applied to the chest wall contour, and the minimum distance between the reference chest wall contour and the transformed chest wall contour was evaluated for each pixel. Results: The measured surface dose for the 2-dot, 6-dot, and 4-dot marker-blocks at 0° incidence were 74%, 71%, and 77% of dose to dmax respectively. At 30° beam incidence this increased to 76%, 72%, and 81%. The best external surface position was patient and fraction dependent, with no consistent best choice. Conclusions: The increase in surface dose directly under the RPM block is approximately equivalent to 3 mm of bolus. No marker-block position on the breast surface was found to be more representative of overall chest wall motion; therefore block positional stability and reproducibility can be used to determine optimal placement on the breast or chest wall.« less

  14. Experimental verification of a two-dimensional respiratory motion compensation system with ultrasound tracking technique in radiation therapy.

    PubMed

    Ting, Lai-Lei; Chuang, Ho-Chiao; Liao, Ai-Ho; Kuo, Chia-Chun; Yu, Hsiao-Wei; Zhou, Yi-Liang; Tien, Der-Chi; Jeng, Shiu-Chen; Chiou, Jeng-Fong

    2018-05-01

    This study proposed respiratory motion compensation system (RMCS) combined with an ultrasound image tracking algorithm (UITA) to compensate for respiration-induced tumor motion during radiotherapy, and to address the problem of inaccurate radiation dose delivery caused by respiratory movement. This study used an ultrasound imaging system to monitor respiratory movements combined with the proposed UITA and RMCS for tracking and compensation of the respiratory motion. Respiratory motion compensation was performed using prerecorded human respiratory motion signals and also sinusoidal signals. A linear accelerator was used to deliver radiation doses to GAFchromic EBT3 dosimetry film, and the conformity index (CI), root-mean-square error, compensation rate (CR), and planning target volume (PTV) were used to evaluate the tracking and compensation performance of the proposed system. Human respiratory pattern signals were captured using the UITA and compensated by the RMCS, which yielded CR values of 34-78%. In addition, the maximum coronal area of the PTV ranged from 85.53 mm 2 to 351.11 mm 2 (uncompensated), which reduced to from 17.72 mm 2 to 66.17 mm 2 after compensation, with an area reduction ratio of up to 90%. In real-time monitoring of the respiration compensation state, the CI values for 85% and 90% isodose areas increased to 0.7 and 0.68, respectively. The proposed UITA and RMCS can reduce the movement of the tracked target relative to the LINAC in radiation therapy, thereby reducing the required size of the PTV margin and increasing the effect of the radiation dose received by the treatment target. Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  15. Computer-aided analysis of star shot films for high-accuracy radiation therapy treatment units

    NASA Astrophysics Data System (ADS)

    Depuydt, Tom; Penne, Rudi; Verellen, Dirk; Hrbacek, Jan; Lang, Stephanie; Leysen, Katrien; Vandevondel, Iwein; Poels, Kenneth; Reynders, Truus; Gevaert, Thierry; Duchateau, Michael; Tournel, Koen; Boussaer, Marlies; Cosentino, Dorian; Garibaldi, Cristina; Solberg, Timothy; De Ridder, Mark

    2012-05-01

    As mechanical stability of radiation therapy treatment devices has gone beyond sub-millimeter levels, there is a rising demand for simple yet highly accurate measurement techniques to support the routine quality control of these devices. A combination of using high-resolution radiosensitive film and computer-aided analysis could provide an answer. One generally known technique is the acquisition of star shot films to determine the mechanical stability of rotations of gantries and the therapeutic beam. With computer-aided analysis, mechanical performance can be quantified as a radiation isocenter radius size. In this work, computer-aided analysis of star shot film is further refined by applying an analytical solution for the smallest intersecting circle problem, in contrast to the gradient optimization approaches used until today. An algorithm is presented and subjected to a performance test using two different types of radiosensitive film, the Kodak EDR2 radiographic film and the ISP EBT2 radiochromic film. Artificial star shots with a priori known radiation isocenter size are used to determine the systematic errors introduced by the digitization of the film and the computer analysis. The estimated uncertainty on the isocenter size measurement with the presented technique was 0.04 mm (2σ) and 0.06 mm (2σ) for radiographic and radiochromic films, respectively. As an application of the technique, a study was conducted to compare the mechanical stability of O-ring gantry systems with C-arm-based gantries. In total ten systems of five different institutions were included in this study and star shots were acquired for gantry, collimator, ring, couch rotations and gantry wobble. It was not possible to draw general conclusions about differences in mechanical performance between O-ring and C-arm gantry systems, mainly due to differences in the beam-MLC alignment procedure accuracy. Nevertheless, the best performing O-ring system in this study, a BrainLab/MHI Vero system

  16. Importance of dosimetry protocol for cell irradiation on a low X-rays facility and consequences for the biological response.

    PubMed

    Dos Santos, Morgane; Paget, Vincent; Ben Kacem, Mariam; Trompier, François; Benadjaoud, Mohamed Amine; François, Agnès; Guipaud, Olivier; Benderitter, Marc; Milliat, Fabien

    2018-06-01

    The main objective of radiobiology is to establish links between doses and radiation-induced biological effects. In this context, well-defined dosimetry protocols are crucial to the determination of experimental protocols. This work proposes a new dosimetry protocol for cell irradiation in a SARRP and shows the importance of the modification of some parameters defined in dosimetry protocol for physical dose and biological outcomes. Once all parameters of the configuration were defined, dosimetry measurements with ionization chambers and EBT3 films were performed to evaluate the dose rate and the attenuation due to the cell culture medium. To evaluate the influence of changes in cell culture volume and/or additional filtration, 6-well plates containing EBT3 films with water were used to determine the impact on the physical dose at 80 kV. Then, experiments with the same irradiation conditions were performed by replacing EBT3 films by HUVECs. The biological response was assessed using clonogenic assay. Using a 0.15 mm copper filter lead to a variation of +1% using medium thickness of 0.104 cm to -8% using a medium thickness of 0.936 cm on the physical dose compare to the reference condition (0.313 cm). For the 1 mm aluminum filter, a variation of +8 to -40% for the same medium thickness conditions has been observed. Cells irradiated in the same conditions showed significant differences in survival fraction, corroborating the effects of dosimetric changes on physical dose. This work shows the importance of dosimetry in radiobiology studies and the need of an accurate description of the dosimetry protocol used for irradiation.

  17. A methodology for dosimetry audit of rotational radiotherapy using a commercial detector array.

    PubMed

    Hussein, Mohammad; Tsang, Yatman; Thomas, Russell A S; Gouldstone, Clare; Maughan, David; Snaith, Julia A D; Bolton, Steven C; Nisbet, Andrew; Clark, Catharine H

    2013-07-01

    To develop a methodology for the use of a commercial detector array in dosimetry audits of rotational radiotherapy. The methodology was developed as part of the development of a national audit of rotational radiotherapy. Ten cancer centres were asked to create a rotational radiotherapy treatment plan for a three-dimensional treatment-planning-system (3DTPS) test and audited. Phantom measurements using a commercial 2D ionisation chamber (IC) array were compared with measurements using 0.125 cm(3) IC, Gafchromic film and alanine pellets in the same plane. Relative and absolute gamma index (γ) comparisons were made for Gafchromic film and 2D-Array planes, respectively. Comparisons between individual detectors within the 2D-Array against the corresponding IC and alanine measurement showed a statistically significant concordance correlation coefficient (both ρc>0.998, p<0.001) with mean difference of -1.1 ± 1.1% and -0.8 ± 1.1%, respectively, in a high dose PTV. In the γ comparison between the 2D-Array and film it was that the 2D-Array was more likely to fail planes where there was a dose discrepancy due to the absolute analysis performed. It has been found that using a commercial detector array for a dosimetry audit of rotational radiotherapy is suitable in place of standard systems of dosimetry. Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.

  18. SU-E-T-771: Two Dimensional Raman Mapping of Carbon Bonds of Radiochromic Films: An Approach to Micro-Dosimetry

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Heo, T; Ye, S

    2015-06-15

    Purpose: To study a feasibility of micro-dosimetry with high dose-sensitivity and resolution using two-dimensional Raman mapping on the basis of carbon bonds concentration of radiochromic films Methods: Unlaminated EBT3 films with the purpose of maximal Raman data acquisition were irradiated by 6 MV beam from 5 MU to 1000 MU at the reference condition. Each film was irradiated with shielding material of lead blocking on the half of film as well as the jaw open in half for distinct dose contrast. Raman peaks of 2070 cm-1, 2095 cm-1, and 2115 cm-1 were major subjects to study, which are assumed tomore » be the spectroscopy of carbon triple bonds of monomers, carbon double bonds of polymers, and carbon triple bonds of polymers, respectively. Laser exposure for Raman spectroscopy generated peak’s trend due to polymerization by laser output and this trend was utilized to find out basic peaks related to polymerization process. The relative dose contrast in each one film was detected by Raman spectroscopy with the aid of an auto-scanning stage, comparing the dose contrast between non-irradiated area and irradiated area. Raman spatial resolution was enhanced up to 20 micrometers, assuming the spatial uniformity of radio¬active rod-shaped LiPCDA crystals. An optical scanner with 9600 dpi was used to scan the red-channel intensity to read the dose contrast for 5 MU delivered film. Results: The peak intensity for Raman wavenumber of 2070 cm-1 was used for mapping since it reflected the different peak intensities based on polymerization degree by irradiation. Dose contrast from 1000MU to 5 MU was distinguished by Raman mapping analysis, whereas optical intensity of red-channel didn’t show any difference. Conclusion: In consideration of laser effect, the quantitative analysis based on raw data of Raman mapping could provide more statistically reliable dosimetry than point measurements.« less

  19. Use of radiochromic film as a high-spatial resolution dosimeter by Raman spectroscopy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Mirza, Jamal Ahmad; Park, Hyeonsuk

    Purpose: Due to increasing demand for high-spatial resolution dosimetry, radiochromic films have been investigated as potential candidates but are often limited by the scanning system, e.g., flatbed optical scanner. In this study, Raman spectroscopy in conjunction with a microscope was selected as an alternative method for high-spatial resolution dosimetry of radiochromic film. Methods: Unlaminated Gafchromic™ EBT3 films were irradiated with doses between 0 and 50 Gy using 6 MV x-rays of a clinical linear accelerator. Depth profiling from the surface of unlaminated film was performed to acquire the maximum Raman intensity peaks of C≡C and C=C stretching bands of diacetylenemore » polymer. The Raman mapping technique for a region of interest (200 × 200, 30 × 30 μm{sup 2}) was developed to reduce a large variation in a Raman spectrum produced with a sampling resolution of a few μm. The preprocessing of Raman spectra was carried out to determine a dosimetric relationship with the amount of diacetylene polymerization. Results: Due to partial diacetylene polymerization upon irradiation, two Raman peaks of C=C and C≡C stretching bands were observed around 1447 and 2060 cm{sup −1}, respectively. The maximum intensities of the two peaks were obtained by positioning a focused laser spot on the surface of unlaminated film. For the dose range of 0–50 Gy, the band heights of both C≡C and C=C peaks increase asymptotically with increasing doses and can be fit with an exponential function of two components. The relative standard deviation in Raman mapping was found to be less than ±5%. By using this technique, dose uniformity was found to be within ±2%. Conclusions: The Raman intensity for C=C and C≡C peaks increases with an increase in the amount of diacetylene polymerization due to an increase in dose. This study shows the potential of Raman spectroscopy as an alternative for absolute dosimetry verifications with a high-spatial resolution of a few μm, but

  20. SU-E-T-299: Dosimetric Characterization of Small Field in Small Animal Irradiator with Radiochromic Films

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Han, S; Kim, K; Jung, H

    Purpose: The small animal irradiator has been used with small animals to optimize new radiation therapy as preclinical studies. The small animal was irradiated by whole- or partial-body exposure. In this study, the dosimetric characterizations of small animal irradiator were carried out in small field using Radiochromic films Material & Methods: The study was performed in commercial animal irradiator (XRAD-320, Precision x-ray Inc, North Brantford) with Radiochromic films (EBT2, Ashland Inc, Covington). The calibration curve was generated between delivery dose and optical density (red channel) and the films were scanned by and Epson 1000XL scanner (Epson America Inc., Long Beach,more » CA).We evaluated dosimetric characterization of irradiator using various filter supported by manufacturer in 260 kV. The various filters were F1 (2.0mm Aluminum (HVL = about 1.0mm Cu) and F2 (0.75mm Tin + 0.25mm Copper + 1.5mm Aluminum (HVL = about 3.7mm Cu). According to collimator size (3, 5, 7, 10 mm, we calculated percentage depth dose (PDD) and the surface –source distance(SSD) was 17.3 cm considering dose rate. Results: The films were irradiated in 260 kV, 10mA and we increased exposure time 5sec. intervals from 5sec. to 120sec. The calibration curve of films was fitted with cubic function. The correlation between optical density and dose was Y=0.1405 X{sup 3}−2.916 X{sup 2}+25.566 x+2.238 (R{sup 2}=0.994). Based on the calibration curve, we calculated PDD in various filters depending on collimator size. When compared PDD of specific depth (3mm) considering animal size, the difference by collimator size was 4.50% in free filter and F1 was 1.53% and F2 was within 2.17%. Conclusion: We calculated PDD curve in small animal irradiator depending on the collimator size and the kind of filter using the radiochromic films. The various PDD curve was acquired and it was possible to irradiate various dose using these curve.« less

  1. Evaluation of detector array technology for the verification of advanced intensity-modulated radiotherapy

    NASA Astrophysics Data System (ADS)

    Hussien, Mohammad

    Purpose: Quality assurance (QA) for intensity modulated radiotherapy (IMRT) has evolved substantially. In recent years, various ionization chamber or diode detector arrays have become commercially available, allowing pre-treatment absolute dose verification with near real-time results. This has led to a wide uptake of this technology to replace point dose and film dosimetry and to facilitate QA streamlining. However, arrays are limited by their spatial resolution giving rise to concerns about their response to clinically relevant deviations. The common factor in all commercial array systems is the reliance on the gamma index (γ) method to provide the quantitative evaluation of the measured dose distribution against the Treatment Planning System (TPS) calculated dose distribution. The mathematical definition of the gamma index presents computational challenges that can cause a variation in the calculation in different systems. The purpose of this thesis was to evaluate the suitability of detector array systems, combined with their implementation of the gamma index, in the verification and dosimetry audit of advanced IMRT. Method: The response of various commercial detector array systems (Delta4®, ArcCHECK®, and the PTW 2D-Array seven29™ and OCTAVIUS II™ phantom combination, Gafchromic® EBT2 and composite EPID measurements) to simulated deliberate changes in clinical IMRT and VMAT plans was evaluated. The variability of the gamma index calculation in the different systems was also evaluated by comparing against a bespoke Matlab-based gamma index analysis software. A novel methodology for using a commercial detector array in a dosimetry audit of rotational radiotherapy was then developed. Comparison was made between measurements using the detector array and those performed using ionization chambers, alanine and radiochromic film. The methodology was developed as part of the development of a national audit of rotational radiotherapy. Ten cancer centres were

  2. SU-E-I-42: Measurement of X-Ray Beam Width and Geometric Efficiency in MDCT Using Radiochromic Films.

    PubMed

    Liillau, T; Liebmann, M; von Boetticher, H; Poppe, B

    2012-06-01

    The purpose of this work was to measure the x-ray beam width and geometric efficiency (GE) of a multi detector computed tomography scanner (MDCT) for different beam collimations using radiochromic films. In MDCT, the primary beam width extends the nominal beam collimation to irradiate the active detector elements uniformly (called 'over-beaming') which contributes to increased radiation dose to the patient compared to single detector CT. Therefore, the precise determination of the primary beam width and GE is of value for any CT dose calculation using Monte Carlo or analytical methods. Single axial dose profiles free in air were measured for 6 different beam collimations nT for a Siemens SOMATOM Sensation 64 Scanner with Gafchromic XR-QA2 films. The films were calibrated relative to the measured charge of a PTW semiflex ionization chamber (type: 31010) for a single rotation in the CT scanner at the largest available beam collimation of 28.8 mm. The beam energy for all measurements in this work was set to 120 kVp. For every measured dose profile and beam collimation the GEin-air and the full-width-at-half- maximum value (FWHM) as a value for the x-ray beam width was determined. Over-beaming factors FWHM / nT were calculated accordingly. For MDCT beam collimations from 7.2 (12×0.6 mm) to 28.8 (24×1.2 mm) the geometric efficiency was between 58 and 85 %. The over- beaming factor ranged from 1.43 to 1.11. For beam collimations of 1×5 mm and 1×10 mm the GE was 77 % and 84 % respectively. The over-beaming factors were close to 1, as expected. This work has shown that radiochromic films can be used for accurate x-ray beam width and geometric efficiency measurements due to their high spatial resolution. The measured free-in-air geometric efficiency and the over-beaming factor depend strongly on beam collimation. © 2012 American Association of Physicists in Medicine.

  3. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Gibbard, G; Khanal, S; Shang, C

    Purpose: To validate a novel real time quality assurance device, as a means to test coincidence of light fields to electron radiation fields. Method and Materials: Use a Raven™ (LAP of America Laser Applications) detector, both light fields and electron radiation fields of various electron cones and cutouts from two Clinacs (Varian Medical) were sensed by the phosphor screen and registered by a CCD camera of the device. During measurements, the screen surface of Raven was placed at ISO center level facing the radiation field at any setup angle. Subsequently the light field and the electron radiation field (with 100more » MU) were captured separately by the device. The measurements were then analyzed using the Raven software with a maximal 25 by 25 cm^2 field size and 0.25 mm resolution. The results were further compared against those using chromic graphic film (EBT3). To ensure consistence only the dimensions through the central axis were recorded. All the films were analyzed with Dose Lab Pro™ v6.70. Coincidence comparisons were done within the tests on one Clinac and between the two Clinacs. Results: The similar difference means of the field coincidence are indicated from the 18 samples by EBT films and 34 by the Raven as 1.33 ± 0.87 mm verses 1.06 ± 0.58 mm (p = 0.74). The mean differences of the field coincidence between the two Clinacs, 1.16 ± 0.68 mm (n=16) and 1.16 ± 0.48 mm (n=18) respectively, also suggest no statistical difference (p= 0.40). Conclusion: The Raven™ is a comparable quality verification device to the chromic graphic film (EBT3) in checking the coincidence between light and electron radiation fields. In this investigation, the Raven device provided highly reproducible results with real time analysis. Further improvements in its resolution and automatic analyzing capability are warranted.« less

  4. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Mendoza-Moctezuma, A. I.; Aguilar, J. Garcia; Garcia-Garduno, O. A.

    Interventional cardiology procedures are an effective alternative for the reestablishment of correct sanguineous circulation in the heart. However, this kind of procedures exposes to the patients to a relatively high radiation doses. Usually, the surface peak skin dose is evaluated using a visual scale with a comparator strip, nevertheless, even if the comparator strip provides a simple and quick method for estimating the dose it has an uncertainty of {+-}25%. For this reason, a better evaluation method is needed. The objective of our project is to determine the surface peak skin dose of interventional cardiology procedures using GafChromic XR-RV2 filmmore » together with a commercial flatbed scanner in reflection mode. Here we report a protocol to handle GafChromic XR-RV2 film using a commercial flat bed scanner in reflection mode aiming at an uncertainty of {+-}3%.« less

  5. Effect of titanium dental implants on proton therapy delivered for head tumors: experimental validation using an anthropomorphic head phantom

    NASA Astrophysics Data System (ADS)

    Oancea, C.; Shipulin, K.; Mytsin, G.; Molokanov, A.; Niculae, D.; Ambrožová, I.; Davídková, M.

    2017-03-01

    A dosimetric experiment was performed at the Medico-Technical Complex in the Joint Institute for Nuclear Research, Dubna, to investigate the effects of metallic dental implants in the treatment of head and neck tumours with proton therapy. The goal of the study was to evaluate the 2D dose distributions of different clinical treatment plans measured in an anthropomorphic phantom, and compare them to predictions from a treatment planning system. The anthropomorphic phantom was sliced into horizontal segments. Two grade 4 Titanium implants were inserted between 2 slices, corresponding to a maxillary area. GafChromic EBT2 films were placed between the segments containing the implants to measure the 2D delivered dose. Two different targets were designed: the first target includes the dental implants in the isocentre, and in the second target, the proton beam is delivered through the implants, which are located at the entrance region of the Bragg curve. The experimental results were compared to the treatment plans made using our custom 3D Treatment Planning System, named RayTreat. To quantitatively determine differences in the isodose distributions (measured and calculated), the gamma index (3 mm, 3%) was calculated for each target for the matrix value in the region of high isodose (> 90%): for the experimental setup, which includes the implants in the SOBP region, the result obtained was 84.3%. When the implants were localised in the entrance region of the Bragg curve, the result obtained was 86.4%. In conclusion, the uncertainties introduced by the clinically planned dose distribution are beyond reasonable limits. The linear energy transfer spectra in close proximity to the implants were investigated using solid state nuclear track detectors (TED). Scattered particles outside the target were detected.

  6. Evaluation of backscatter dose from internal lead shielding in clinical electron beams using EGSnrc Monte Carlo simulations

    PubMed Central

    Marsh, Steven

    2015-01-01

    Internal lead shielding is utilized during superficial electron beam treatments of the head and neck, such as lip carcinoma. Methods for predicting backscattered dose include the use of empirical equations or performing physical measurements. The accuracy of these empirical equations required verification for the local electron beams. In this study, a Monte Carlo model of a Siemens Artiste linac was developed for 6, 9, 12, and 15 MeV electron beams using the EGSnrc MC package. The model was verified against physical measurements to an accuracy of better than 2% and 2 mm. Multiple MC simulations of lead interfaces at different depths, corresponding to mean electron energies in the range of 0.2–14 MeV at the interfaces, were performed to calculate electron backscatter values. The simulated electron backscatter was compared with current empirical equations to ascertain their accuracy. The major finding was that the current set of backscatter equations does not accurately predict electron backscatter, particularly in the lower energies region. A new equation was derived which enables estimation of electron backscatter factor at any depth upstream from the interface for the local treatment machines. The derived equation agreed to within 1.5% of the MC simulated electron backscatter at the lead interface and upstream positions. Verification of the equation was performed by comparing to measurements of the electron backscatter factor using Gafchromic EBT2 film. These results show a mean value of 0.997±0.022 to 1σ of the predicted values of electron backscatter. The new empirical equation presented can accurately estimate electron backscatter factor from lead shielding in the range of 0.2 to 14 MeV for the local linacs. PACS numbers: 87.53.Bn, 87.55.K‐, 87.56.bd PMID:26699566

  7. SU-E-T-373: Evaluation and Reduction of Contralateral Skin /subcutaneous Dose for Tangential Breast Irradiation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Butson, M; Carroll, S; Whitaker, M

    2015-06-15

    Purpose: Tangential breast irradiation is a standard treatment technique for breast cancer therapy. One aspect of dose delivery includes dose delivered to the skin caused by electron contamination. This effect is especially important for highly oblique beams used on the medical tangent where the electron contamination deposits dose on the contralateral breast side. This work aims to investigate and predict as well as define a method to reduce this dose during tangential breast radiotherapy. Methods: Analysis and calculation of breast skin and subcutaneous dose is performed using a Varian Eclipse planning system, AAA algorithm for 6MV x-ray treatments. Measurements weremore » made using EBT3 Gafchromic film to verify the accuracy of planning data. Various materials were tested to assess their ability to remove electron contamination on the contralateral breast. Results: Results showed that the Varian Eclipse AAA algorithm could accurately estimate contralateral breast dose in the build-up region at depths of 2mm or deeper. Surface dose was underestimated by the AAA algorithm. Doses up to 12% of applied dose were seen on the contralateral breast surface and up to 9 % at 2mm depth. Due to the nature of this radiation, being mainly low energy electron contamination, a bolus material could be used to reduce this dose to less than 3%. This is accomplished by 10 mm of superflab bolus or by 1 mm of lead. Conclusion: Contralateral breast skin and subcutaneous dose is present for tangential breast treatment and has been measured to be up to 12% of applied dose from the medial tangent beam. This dose is deposited at shallow depths and is accurately calculated by the Eclipse AAA algorithm at depths of 2mm or greater. Bolus material placed over the contralateral can be used to effectively reduce this skin dose.« less

  8. Comparison of trigeminal neuralgia radiosurgery plans using two film detectors for the commissioning of small photon beams

    PubMed Central

    Esparza‐Moreno, Karina P.; Ballesteros‐Zebadúa, Paola; Lárraga‐Gutiérrez, José M.; Moreno‐Jiménez, Sergio; Celis‐López, Miguel A.

    2013-01-01

    Trigeminal neuralgia (TN) is a chronic, episodic facial pain syndrome that can be extremely intense, and it occurs within the regions of the face that are innervated by the three branches of the trigeminal nerve. Stereotactic radiosurgery (SRS) is the least invasive procedure to treat TN. SRS uses narrow photon beams that require high spatial resolution techniques for their measurement. The use of radiographic or radiochromic films for small‐field dosimetry is advantageous because high spatial resolution and two‐dimensional dose measurements can be performed. Because these films have different properties, it is expected that the calculated dose distributions for TN patients will behave differently, depending on the detector used for the commissioning of the small photon beams. This work is based on two sets of commissioned data: one commissioned with X‐OMAT V2 film and one commissioned with EBT2 film. The calculated dose distributions for 23 TN patients were compared between the commissioning datasets. The variables observed were the differences in the half widths of the 35 and 40 Gy isodose lines (related to the entrance distance to the brainstem) and the volume of the brainstem that received a dose of 12 Gy or more (V12). The results of this comparison showed that there were statistically significant differences between the two calculated dose distributions. The magnitudes of these differences were up to 0.33 mm and 0.38 mm for the 35 and 40 Gy isodose lines. The corresponding difference for the V12 was up to 2.1 cc. It is clear that these differences may impact the treatment of TN patients, and then it must be important to perform this type of analysis when observing complication rates. Clinical reports on irradiation techniques for trigeminal neuralgia should consider that different detectors used for commissioning treatment planning systems might result in small but significant differences in dose distributions. PACS number: 87.55.km PMID:24257267

  9. SU-E-J-101: Retroactive Calculation of TLD and Film Dose in Anthropomorphic Phantom as Assessment of Updated TPS Performance

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Alkhatib, H; Oves, S

    Purpose: To demonstrate a quick and comprehensive method verifying the accuracy of the updated dose model by recalculating dose distribution in an anthropomorphic phantom with a new version of the TPS and comparing the results to measured values. Methods: CT images and IMRT plan of an RPC anthropomorphic head phantom, previously calculated by Pinnacle 9.0, was re-computed using Pinnacle 9.2 and 9.6. The dosimeters within the phantom include four TLD capsules representing a primary PTV, two TLD capsules representing a secondary PTV, and two TLD capsules representing an organ at risk. Also included were three sheets of Gafchromic film. Performancemore » of the updated TPS version was assessed by recalculating point doses and dose profiles corresponding to TLD and film position respectively and then comparing the results to reported values by the RPC. Results: Comparing calculated doses to reported measured doses from the RPC yielded an average disagreement of 1.48%, 2.04% and 2.10% for versions 9.0, 9.2, 9.6 respectively. Computed doses points all meet the RPC's passing criteria with the exception of the point representing the superior organ at risk in version 9.6. However, qualitative analysis of the recalculated dose profiles showed improved agreement with those of the RPC, especially in the penumbra region. Conclusion: This work has demonstrated the calculation results of Pinnacle 9.2 and 9.6 vs 9.0 version. Additionally, this study illustrates a method for the user to gain confidence upgrade to a newer version of the treatment planning system.« less

  10. A critical evaluation of the PTW 2D‐ARRAY seven29 and OCTAVIUS II phantom for IMRT and VMAT verification

    PubMed Central

    Adams, Elizabeth J.; Jordan, Thomas J.; Clark, Catharine H.; Nisbet, Andrew

    2013-01-01

    Quality assurance (QA) for intensity‐ and volumetric‐modulated radiotherapy (IMRT and VMAT) has evolved substantially. In recent years, various commercial 2D and 3D ionization chamber or diode detector arrays have become available, allowing for absolute verification with near real time results, allowing for streamlined QA. However, detector arrays are limited by their resolution, giving rise to concerns about their sensitivity to errors. Understanding the limitations of these devices is therefore critical. In this study, the sensitivity and resolution of the PTW 2D‐ARRAY seven29 and OCTAVIUS II phantom combination was comprehensively characterized for use in dynamic sliding window IMRT and RapidArc verification. Measurement comparisons were made between single acquisition and a multiple merged acquisition techniques to improve the effective resolution of the 2D‐ARRAY, as well as comparisons against GAFCHROMIC EBT2 film and electronic portal imaging dosimetry (EPID). The sensitivity and resolution of the 2D‐ARRAY was tested using two gantry angle 0° modulated test fields. Deliberate multileaf collimator (MLC) errors of 1, 2, and 5 mm and collimator rotation errors were inserted into IMRT and RapidArc plans for pelvis and head & neck sites, to test sensitivity to errors. The radiobiological impact of these errors was assessed to determine the gamma index passing criteria to be used with the 2D‐ARRAY to detect clinically relevant errors. For gamma index distributions, it was found that the 2D‐ARRAY in single acquisition mode was comparable to multiple acquisition modes, as well as film and EPID. It was found that the commonly used gamma index criteria of 3% dose difference or 3 mm distance to agreement may potentially mask clinically relevant errors. Gamma index criteria of 3%/2 mm with a passing threshold of 98%, or 2%/2 mm with a passing threshold of 95%, were found to be more sensitive. We suggest that the gamma index passing thresholds may be used

  11. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Paudel, M R; Beachey, D J; Sarfehnia, A

    Purpose: A new commercial GPU-based Monte Carlo dose calculation algorithm (GPUMCD) developed by the vendor Elekta™ to be used in the Monaco Treatment Planning System (TPS) is capable of modeling dose for both a standard linear accelerator and for an Elekta MRI-Linear accelerator (modeling magnetic field effects). We are evaluating this algorithm in two parts: commissioning the algorithm for an Elekta Agility linear accelerator (the focus of this work) and evaluating the algorithm’s ability to model magnetic field effects for an MRI-linear accelerator. Methods: A beam model was developed in the Monaco TPS (v.5.09.06) using the commissioned beam data formore » a 6MV Agility linac. A heterogeneous phantom representing tumor-in-lung, lung, bone-in-tissue, and prosthetic was designed/built. Dose calculations in Monaco were done using the current clinical algorithm (XVMC) and the new GPUMCD algorithm (1 mm3 voxel size, 0.5% statistical uncertainty) and in the Pinnacle TPS using the collapsed cone convolution (CCC) algorithm. These were compared with the measured doses using an ionization chamber (A1SL) and Gafchromic EBT3 films for 2×2 cm{sup 2}, 5×5 cm{sup 2}, and 10×10 cm{sup 2} field sizes. Results: The calculated central axis percentage depth doses (PDDs) in homogeneous solid water were within 2% compared to measurements for XVMC and GPUMCD. For tumor-in-lung and lung phantoms, doses calculated by all of the algorithms were within the experimental uncertainty of the measurements (±2% in the homogeneous phantom and ±3% for the tumor-in-lung or lung phantoms), except for 2×2 cm{sup 2} field size where only the CCC algorithm differs from film by 5% in the lung region. The analysis for bone-in-tissue and the prosthetic phantoms are ongoing. Conclusion: The new GPUMCD algorithm calculated dose comparable to both the XVMC algorithm and to measurements in both a homogeneous solid water medium and the heterogeneous phantom representing lung or tumor-in-lung for

  12. SU-E-T-472: A Multi-Dimensional Measurements Comparison to Analyze a 3D Patient Specific QA Tool

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ashmeg, S; Jackson, J; Zhang, Y

    Purpose: To quantitatively evaluate a 3D patient specific QA tool using 2D film and 3D Presage dosimetry. Methods: A brain IMRT case was delivered to Delta4, EBT2 film and Presage plastic dosimeter. The film was inserted in the solid water slabs at 7.5cm depth for measurement. The Presage dosimeter was inserted into a head phantom for 3D dose measurement. Delta4's Anatomy software was used to calculate the corresponding dose to the film in solid water slabs and to Presage in the head phantom. The results from Anatomy were compared to both calculated results from Eclipse and measured dose from filmmore » and Presage to evaluate its accuracy. Using RIT software, we compared the “Anatomy” dose to the EBT2 film measurement and the film measurement to ECLIPSE calculation. For 3D analysis, DICOM file of “Anatomy” was extracted and imported to CERR software, which was used to compare the Presage dose to both “Anatomy” calculation and ECLIPSE calculation. Gamma criteria of 3% - 3mm and 5% - 5mm was used for comparison. Results: Gamma passing rates of film vs “Anatomy”, “Anatomy” vs ECLIPSE and film vs ECLIPSE were 82.8%, 70.9% and 87.6% respectively when 3% - 3mm criteria is used. When the criteria is changed to 5% - 5mm, the passing rates became 87.8%, 76.3% and 90.8% respectively. For 3D analysis, Anatomy vs ECLIPSE showed gamma passing rate of 86.4% and 93.3% for 3% - 3mm and 5% - 5mm respectively. The rate is 77.0% for Presage vs ECLIPSE analysis. The Anatomy vs ECLIPSE were absolute dose comparison. However, film and Presage analysis were relative comparison Conclusion: The results show higher passing rate in 3D than 2D in “Anatomy” software. This could be due to the higher degrees of freedom in 3D than in 2D for gamma analysis.« less

  13. Validation of GPU-accelerated superposition-convolution dose computations for the Small Animal Radiation Research Platform.

    PubMed

    Cho, Nathan; Tsiamas, Panagiotis; Velarde, Esteban; Tryggestad, Erik; Jacques, Robert; Berbeco, Ross; McNutt, Todd; Kazanzides, Peter; Wong, John

    2018-05-01

    The Small Animal Radiation Research Platform (SARRP) has been developed for conformal microirradiation with on-board cone beam CT (CBCT) guidance. The graphics processing unit (GPU)-accelerated Superposition-Convolution (SC) method for dose computation has been integrated into the treatment planning system (TPS) for SARRP. This paper describes the validation of the SC method for the kilovoltage energy by comparing with EBT2 film measurements and Monte Carlo (MC) simulations. MC data were simulated by EGSnrc code with 3 × 10 8 -1.5 × 10 9 histories, while 21 photon energy bins were used to model the 220 kVp x-rays in the SC method. Various types of phantoms including plastic water, cork, graphite, and aluminum were used to encompass the range of densities of mouse organs. For the comparison, percentage depth dose (PDD) of SC, MC, and film measurements were analyzed. Cross beam (x,y) dosimetric profiles of SC and film measurements are also presented. Correction factors (CFz) to convert SC to MC dose-to-medium are derived from the SC and MC simulations in homogeneous phantoms of aluminum and graphite to improve the estimation. The SC method produces dose values that are within 5% of film measurements and MC simulations in the flat regions of the profile. The dose is less accurate at the edges, due to factors such as geometric uncertainties of film placement and difference in dose calculation grids. The GPU-accelerated Superposition-Convolution dose computation method was successfully validated with EBT2 film measurements and MC calculations. The SC method offers much faster computation speed than MC and provides calculations of both dose-to-water in medium and dose-to-medium in medium. © 2018 American Association of Physicists in Medicine.

  14. Characterization of XR-RV3 GafChromic{sup ®} films in standard laboratory and in clinical conditions and means to evaluate uncertainties and reduce errors

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Farah, J., E-mail: jad.farah@irsn.fr; Clairand, I.; Huet, C.

    2015-07-15

    Purpose: To investigate the optimal use of XR-RV3 GafChromic{sup ®} films to assess patient skin dose in interventional radiology while addressing the means to reduce uncertainties in dose assessment. Methods: XR-Type R GafChromic films have been shown to represent the most efficient and suitable solution to determine patient skin dose in interventional procedures. As film dosimetry can be associated with high uncertainty, this paper presents the EURADOS WG 12 initiative to carry out a comprehensive study of film characteristics with a multisite approach. The considered sources of uncertainties include scanner, film, and fitting-related errors. The work focused on studying filmmore » behavior with clinical high-dose-rate pulsed beams (previously unavailable in the literature) together with reference standard laboratory beams. Results: First, the performance analysis of six different scanner models has shown that scan uniformity perpendicular to the lamp motion axis and that long term stability are the main sources of scanner-related uncertainties. These could induce errors of up to 7% on the film readings unless regularly checked and corrected. Typically, scan uniformity correction matrices and reading normalization to the scanner-specific and daily background reading should be done. In addition, the analysis on multiple film batches has shown that XR-RV3 films have generally good uniformity within one batch (<1.5%), require 24 h to stabilize after the irradiation and their response is roughly independent of dose rate (<5%). However, XR-RV3 films showed large variations (up to 15%) with radiation quality both in standard laboratory and in clinical conditions. As such, and prior to conducting patient skin dose measurements, it is mandatory to choose the appropriate calibration beam quality depending on the characteristics of the x-ray systems that will be used clinically. In addition, yellow side film irradiations should be preferentially used since they showed a

  15. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Huang, M; Brezovich, I; Duan, J

    Purpose: To demonstrate a patient specific, image-guided quality assurance method that tests both dosimetric and geometric accuracy for single-isocenter multiple-target VMAT radiosurgery (SIMT-VMAT-SRS) Method: We used a new film type, EBT-XD (optimal range 0.4–40Gy), and an in-house PMMA phantom having a coronal plane for film and a 0.125 cm3 ionization chamber (IC). The phantom contained fiducial features for kV image guided setup and for accurate film marking. Five patient plans with multiple targets sizes ranging from 3 to 21mm in diameter and prescribed doses from 14 to 18 Gy were selected. Two verification plans were generated for each case withmore » the film plane passing through the center of the largest and smallest targets. For the four largest targets we obtained an IC measurement. For each case, a calibration film was irradiated using a custom designed step pattern. The films were scanned using a flatbed color scanner and converted to dose using the calibration film and the three channel calibration method. Image registration was performed between film and treatment planning system calculations to evaluate the geometric accuracy. Results: The mean registration vector had an average magnitude of 0.47 mm (range from 0.13mm to 0.64 mm). For the four largest targets, the mean ratio of the IC and film measurement to expected dose was 0.990 (range 0.968 to 1.009) and 1.032 (1.021 to 1.046), respectively. The fraction of pixels having gamma index < 1 for criteria of 3%/3mm, 3%/2mm, 3%/1mm was 98.8%, 97.5% and 87.2% before geometric registration and 99.1%, 98.3% and 94.8% after registration. Conclusion: We have demonstrated an image-guided QA method can assess both geometric and dosimetric accuracy. The phantom was positioned with sub-millimeter accuracy. Absolute film dosimetry using EBT-XD film was sufficiently accurate for assessment of dose to multi-targets too small for IC measurement in SRS VMAT plans.« less

  16. TU-D-209-06: Head and Neck Tissue Dose From X-Ray Scatter to Physicians Performing Cardiovascular Procedures

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fetterly, K; Schueler, B; Grams, M

    Purpose: The purpose of this work was to characterize the spatial distribution of scatter radiation to the head and neck of a physician performing an x-ray interventional procedure and assess brain, eye lens, and carotid artery dose. Methods: Radiographic x-ray beams were tuned to match the peak energy (56 to 106 keV) and HVL (3.5 to 6.5 mm Al) of x-ray scatter originating from a patient during a fluoroscopic procedure. The radiographic beam was directed upon a Rando phantom from an inferior-left location to mimic a typical patient-operator geometric relationship. A lead-equivalent protective garment was secured to the phantom. Directmore » exposure Gafchromic film (XRQA2) was placed between the transverse plane layers of the head and neck region of the phantom and exposed with 4 scatter-equivalent radiographic beams. A 3×3 cm{sup 2} film placed at the left collar of the phantom was used to monitor incident dose in the position of a radiation monitoring badge. The films were converted to 2D dose distribution maps using FilmQA Pro software and an Epson 11000-XL scanner. The 2D dose distributions maps were normalized by the left collar dose and the percent of left collar dose (%LCD) was calculated for select tissues. Results: The dose maps had high dynamic range (10{sub 4}) and spatial detail. Considering all transverse planes and 4 scatter beam qualities, the median %LCD values were: whole brain 8.5%, left brain 13%, right brain 5.4%, left eye lens 67%, right eye lens 25%, left carotid artery 72%, and right carotid artery 28%. Conclusion: Scatter radiation dose to an operator can be simulated using a tuned radiographic beam and used to expose a phantom and Gafchromic film, thereby creating detailed 2D dose distribution maps. This work facilitates individualized estimation of dose to select head and neck tissues based on an operator’s radiation monitoring badge value.« less

  17. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lee, N

    Purpose: Ocular proton therapy has the following advantages: i) sparing optic nerve, ii) the minimal dose is delivered to surrounding normal tissues. Since the proton therapy center was opened in 2007, 30 patients with ocular tumor have been treated at National Cancer Center using single scattering technique. To develop a solid eye phantom which can verify the output and the beam range with EBT3 film for independent patient QA in ocular proton therapy. Methods: The proton therapy is very effective to treat ocular tumor, because of the Bragg peak feature. In general, the beam shape of eye treatment is aboutmore » 3 cm in diameter and the beam range is under 5 cm. However, proton therapy has uncertainty of beam range problem due to various stopping power of normal tissue, bone, air and so on, so that we should verify the beam range before treatment. For this purpose, a new PMMA phantom with wedge has been developed to use the film dosimetry and the ionization chamber. It is able to place a film on the slope of the phantom, which the spread out Bragg Peak by the water equivalent thickness value of PMMA can be made on the film. We considered to relation with quenching effect of proton energy and range for simple second check. In addition, the ionization chamber (Pin-point chamber, PTW 31014) can be inserted into a hole in the phantom to measure the absolute dose. Results: The output difference and beam range difference were less than 2% and 1.0 mm, respectively, between the measurement and the plan. Conclusion: An eye phantom was developed and its performance was evaluated successfully and it was useful to verify the output. Also with EBT3 film with the quenching effect for measurement depth-dose profile, range to patient QA.« less

  18. SU-E-T-541: Bolus Effect of Thermoplastic Masks in IMRT and VMAT Head and Neck Treatments

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zhen, H; Nedzi, L; Chen, S

    2014-06-01

    Purpose: To quantitatively evaluate the bolus effect of thermoplalstic mask on patient skin dose during multi-field IMRT and VMAT treatment. Methods: The clinically approved target contours for five head and neck patients were deformably registered to an anthropomorphic Rando phantom. Two plans: Multifield IMRT plan with 7-9 beams and VMAT plan with 2-4 arcs were created for each patient following same dose constraints. 3mm skin was excluded from PTVs but not constrained during optimization. The prescription dose was 200-220 cGy/fraction. A thermoplastic head and shoulder mask was customized for the Rando phantom. Each plan was delivered to the phantom twicemore » with and without mask. During each delivery, two rectangular strips of EBT3 films (1cm x 6.8cm) were placed across the anterior upper and lower neck near PTVs to measure the surface dose. For consistency films were positioned at same locations for same patient. A total of 8 film strips were obtained for each patient. Film dose was calibrated in the range of 0-400cGy on the day of plan delivery. For dose comparison 3 regions of interests (ROIs) of 1×1 cm{sup 2} were selected at left, right and middle part of each film, resulting in 6 point doses at each plan delivery. Results: The films without mask show relatively uniform dose distribution while those with mask clearly show mesh pattern of mask, usually indicating an increase in skin dose. On average the increase in skin dose over all ROIs with mask was 31.9%(±14.8%) with a range of 11.4%- 58.4%. There is no statistically significant difference (p=0.44) between skin dose increase in VMAT (30.8%±15.3%) and IMRT delivery (33.0%±14.9%). Conclusion: Thermoplastic immobilization masks increase surface dose for HN patient by around 30%. The magnitude is comparable between multi-field IMRT and VMAT. Radiochromic EBT3 film serves as an effective tool to quantify bolus effect.« less

  19. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kalavagunta, C; Lin, M; Snider, J

    Purpose: To quantify the factors leading to thermoplastic mask bolus-associated-increased skin dose in head and neck IMRT/VMAT using EBT2 film. Methods: EBT2 film placed beneath a dual layer 3-point ORFIT head, neck and shoulder mask was used to test the effect of mask thickness, beam modulation, air gap, and beam obliquity on bolus effect. Mask thickness was varied based on the distribution of 1.6mm Orfilight layer on top of 2 mm Efficast layer. Beam modulation was varied by irradiating the film with an open field (no beam modulation) and a step and shoot field (beam modulation). Air gap between maskmore » and film was varied from 0 to 5mm. Beam obliquity was varied by irradiating the film at gantry angles of 0°, 35°, and 70°.Finally, film strips placed on a Rando phantom under an Orfit mask, in regions of expected high dose, were irradiated using 5 IMRT and 5 VMAT plans with various modulation levels (modulation factor 2 to 5) and the results were compared with those obtained placing OSLDs at the same locations. Results: An 18–34% increase in mask bolus effect was observed for three factors where the effect of beam obliquity ≥ beam modulation > mask thickness. No increase in mask bolus effect was observed for change in air gap. A 6–13% increase in dose due to mask bolus effect was observed on film strips. Conclusion: This work underlines the role of beam obliquity and beam modulation combined with thermoplastic mask thickness in increasing mask bolus-associated skin dose in head and neck IMRT/VMAT. One possible method of dose reduction, based on knowledge gained from this work, is inclusion of skin as an avoidance structure in treatment planning. Another approach is to design a mask with the least amount of thermoplastic material necessary for immobilization.« less

  20. Unintended Consequences of Evidence-Based Treatment Policy Reform: Is Implementation the Goal or the Strategy for Higher Quality Care?

    PubMed

    Park, Alayna L; Tsai, Katherine H; Guan, Karen; Chorpita, Bruce F

    2018-02-14

    This study examined patterns of evidence-based treatment (EBT) delivery following a county-wide EBT reform initiative. Data were gathered from 60 youth and their 21 providers, who were instructed to deliver therapy as they normally would under the EBT initiative. Results showed limited applicability of county-supported EBTs to this service sample, and that most youth did not receive traditional delivery of EBTs. Findings suggest that it may be unrealistic to expect providers to deliver EBTs with fidelity with all clients, and that EBT implementation may be best thought of as a strategy for improving mental health services rather than a goal.

  1. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Mao, R; Tian, L; Ge, H

    Purpose: To evaluate the dosimetry of microscopic disease (MD) region of lung cancer in stereotactic body radiation therapy (SBRT). Methods: For simplicity, we assume organ moves along one dimension. The probability distribution function of tumor position was calculated according to the breathing cycle. The dose to the MD region was obtained through accumulating the treatment planning system calculated doses at different positions in a breathing cycle. A phantom experiment was then conducted to validate the calculated results using a motion phantom (The CIRS ‘Dynamic’ Thorax Phantom). The simulated breathing pattern used a cos4(x) curve with an amplitude of 10mm. Amore » 3-D conformal 7-field plan with 6X energy was created and the dose was calculated in the average intensity projection (AIP) simulation CT images. Both films (EBT2) and optically stimulated luminescence (OSL) detectors were inserted in the target of the phantom to measure the dose during radiation delivery (Varian Truebeam) and results were compared to planning dose parameters. Results: The Gamma analysis (3%/3mm) between measured dose using EBT2 film and calculated dose using AIP was 80.5%, indicating substantial dosimetric differences. While the Gamma analysis (3%/3mm) between measured dose using EBT2 and accumulated dose using 4D-CT was 98.9%, indicating the necessity of dose accumulation using 4D-CT. The measured doses using OSL and theoretically calculated doses using probability distribution function at the corresponding position were comparable. Conclusion: Use of static dose calculation in the treatment planning system could substantially underestimate the actually delivered dose in the MD region for a moving target. Funding Supported by NSFC, No.81372436.« less

  2. Research, Data, and Evidence-Based Treatment Use in State Behavioral Health Systems, 2001-2012

    PubMed Central

    Bruns, Eric J.; Kerns, Suzanne E.U.; Pullmann, Michael D.; Hensley, Spencer W.; Lutterman, Ted; Hoagwood, Kimberly

    2016-01-01

    Objectives Empirical study of public behavioral health systems' use of data and their investment in evidence based treatment (EBT) is limited. This study describes trends in state-level EBT investment and research supports from 2001-2012. Methods Multilevel models examined change over time related to state adoption of EBTs, numbers served, and penetration rates for six behavioral health EBTs for adults and children, based on data from the National Association for State Mental Health Program Directors Research Institute (NRI). State supports related to research, evaluation, and information management were also examined. Results Increasing percentages of states reported funding an external research center, promoting the adoption of EBTs through provider contracts, and providing financial incentives for EBTs. Decreasing percentages of states reported promoting EBT adoption through stakeholder mobilization, monitoring fidelity, and specific budget requests. There was greater reported use of adult-focused EBTs (65-80%) compared with youth EBTs (25-50%). Overall penetration rates of EBTs were low (1% - 3%) and EBT adoption by states showed flat or declining trends. SMHAs' investment in data systems and use of research showed little change. Conclusion SMHA investment in EBTs, implementation infrastructure, and use of research has declined. More systematic measurement and examination of these metrics may provide a useful approach for setting priorities, evaluating success of health reform efforts, and making future investments. PMID:26695495

  3. Technical Note: Evaluation of plastic scintillator detector for small field stereotactic patient-specific quality assurance.

    PubMed

    Qin, Yujiao; Gardner, Stephen J; Kim, Joshua; Huang, Yimei; Wen, Ning; Doemer, Anthony; Chetty, Indrin J

    2017-10-01

    To evaluate the performance of a commercial plastic scintillator detector (PSD) for small-field stereotactic patient-specific quality assurance (QA) measurements using flattening-filter-free beam. A total of 10 spherical targets [volume range: (0.03 cc-2 cc)] were planned with two techniques: (a) dynamic conformal arc (DCA-10 plans) and (b) volumetric modulated arc therapy (VMAT-10 plans). All plans were generated using Varian Eclipse treatment planning system, and AcurosXB v.13 algorithm in 1.0 mm grid size. Additionally, 14 previously treated cranial and spine SRS plans were evaluated [6 DCA, 8 VMAT, volume range: (0.04 cc-119.02 cc)]. Plan modulation was quantified via two metrics: MU per prescription dose (MU/Rx) and Average Leaf Pair Opening (ALPO). QA was performed on the Varian Edge linear accelerator equipped with HDMLC. Three detectors were used: (a) PinPoint ion chamber (PTW; active volume 0.015 cc), (b) Exradin W1 PSD (Standard Imaging; active volume 0.002 cc), and (c) Gafchromic EBT3 film (Ashland). PinPoint chamber and PSD were positioned perpendicular to beam axis in a Lucy phantom (Standard Imaging); films were placed horizontally capturing the coronal plane. PSD, film, and PinPoint chamber measured average differences of 1.00 ± 1.54%, 1.30 ± 1.69%, and -0.66 ± 2.36%, respectively, compared to AcurosXB dose calculation. As the target volume decreased, PinPoint chamber measured lower doses (maximum -5.07% at 0.07 cc target), while PSD and film measured higher doses (2.87% and 2.54% at 0.03 cc target) than AcurosXB. Film agreed with the benchmark detector PSD by an average difference of 0.31 ± 1.20%, but suffered from larger uncertainty; PinPoint chamber underestimated dose by more than 4% for targets smaller than 0.2 cc. Taking PSD as the measurement standard, DCA plans achieved good QA results across all volumes studied, with an average of -0.07 ± 0.89%; for VMAT plans, PSD measured consistently higher dose (1.95 ± 1

  4. Use of a novel extended blink test to evaluate the performance of two polyvinylpyrrolidone-containing, silicone hydrogel contact lenses

    PubMed Central

    Schafer, Jeffery; Reindel, William; Steffen, Robert; Mosehauer, Gary; Chinn, Joseph

    2018-01-01

    Background Sustained digital display viewing reduces eye blink frequency and tear film stability. To retain water and preserve a smooth optical surface, contact lens manufacturers have integrated the humectant polyvinylpyrrolidone (PVP) into silicone hydrogel contact lenses. In this study, extended blink time (EBT) was used to assess visual stability over a prolonged blink interval of two PVP-containing silicone hydrogel lenses, samfilcon A (SAM) and senofilcon A (SEN). Materials and methods This randomized, bilateral, masked, crossover study assessed lens performance in ten subjects after 16 hours of wear. EBT, ie, the time elapsed between cessation of blinking and blur-out of a threshold letter on the acuity chart, was measured. At the end of the wear period, subjects reported duration of computer use and rated visual quality (VQ) and comfort while wearing the assigned lens, and the investigator evaluated lens surface wetting characteristics. Each lens was removed and immediately weighed to determine total water content. Results EBTs were 10.42 seconds for SAM and 8.04 seconds for SEN (p = 0.015). Subjective ratings of VQ after 16 hours of wear were 84.6 for SAM and 74.4 for SEN (p = 0.049). Comfort ratings were 85.9 for SAM and 80.2 for SEN (p > 0.05). Median times of computer use were 6–8 hours for both lens types. Post blink, 70.0% of SAM and 30.0% of SEN lenses were completely wet (p = 0.021). Total water content after wear was 43.7% for SAM and 35.5% for SEN (p < 0.001). Conclusion EBT measurement indicated more stable vision with the PVP-containing SAM polymer compared with the PVP-containing SEN polymer. The SAM polymer also demonstrated better surface wetting and maintained higher water content after a prolonged period of wear. EBT can be valuable in assessing vision stability of patients after hours of computer use. PMID:29765195

  5. SU-F-SPS-10: The Dosimetric Comparison of GammaKnife and Cyberknife Treatment Plans for Brain SRS Treatment

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sanli, E; Mabhouti, H; Cebe, M

    Purpose: Brain stereotactic radiosurgery (SRS) involves the use of precisely directed, single session radiation to create a desired radiobiologic response within the brain target with acceptable minimal effects on surrounding structures or tissues. In this study, the dosimetric comparison of GammaKnife perfection and Cyberknife M6 treatment plans were made. Methods: Treatment plannings were done for GammaKnife perfection unit using Gammaplan treatment planning system (TPS) on the CT scan of head and neck randophantom simulating the treatment of sterotactic treatments for one brain metastasis. The dose distribution were calculated using TMR 10 algorithm. The treatment planning for the same target weremore » also done for Cyberknife M6 machine using Multiplan (TPS) with Monte Carlo algorithm. Using the same film batch, the net OD to dose calibration curve was obtained using both machine by delivering 0- 800 cGy. Films were scanned 48 hours after irradiation using an Epson 1000XL flatbed scanner. Dose distribution were measured using EBT3 film dosimeter. The measured and calculated doses were compared. Results: The dose distribution in the target and 2 cm beyond the target edge were calculated on TPSs and measured using EBT3 film. For cyberknife treatment plans, the gamma analysis passing rates between measured and calculated dose distributions were 99.2% and 96.7% for target and peripheral region of target respectively. For gammaknife treatment plans, the gamma analysis passing rates were 98.9% and 93.2% for target and peripheral region of target respectively. Conclusion: The study shows that dosimetrically comparable plans are achievable with Cyberknife and GammaKnife. Although TMR 10 algorithm predicts the target dose.« less

  6. Research, Data, and Evidence-Based Treatment Use in State Behavioral Health Systems, 2001-2012.

    PubMed

    Bruns, Eric J; Kerns, Suzanne E U; Pullmann, Michael D; Hensley, Spencer W; Lutterman, Ted; Hoagwood, Kimberly E

    2016-05-01

    Empirical study of public behavioral health systems' use of data and their investment in evidence-based treatments (EBTs) is limited. This study describes trends in state-level EBT investment and research supports from 2001 to 2012. Data were from National Association for State Mental Health Program Directors Research Institute (NRI) surveys, which were completed by representatives of state mental health authorities (SMHAs). Multilevel models examined change over time related to state adoption of EBTs, numbers of clients served, and penetration rates for six behavioral health EBTs for adults and children: supported housing, supported employment, assertive community treatment, therapeutic foster care, multisystemic therapy, and functional family therapy. State supports related to research, evaluation, and information management were also examined. Increasing percentages of states reported funding an external research center, promoting the adoption of EBTs through provider contracts, and providing financial incentives for EBTs. Decreasing percentages of states reported promoting EBT adoption through stakeholder mobilization, monitoring fidelity, and specific budget requests. There was greater reported use of adult-focused EBTs (65%-80%) compared with youth-focused EBTs (25%-50%). Overall penetration rates of EBTs were low (1%-3%) and EBT adoption by states showed flat or declining trends. SMHAs' investment in data systems and use of research showed little change. SMHA investment in EBTs, implementation infrastructure, and use of research has declined. More systematic measurement and examination of these metrics may provide a useful approach for setting priorities, evaluating success of health reform efforts, and making future investments.

  7. SU-E-T-344: Validation and Clinical Experience of Eclipse Electron Monte Carlo Algorithm (EMC)

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Pokharel, S; Rana, S

    2014-06-01

    Purpose: The purpose of this study is to validate Eclipse Electron Monte Carlo (Algorithm for routine clinical uses. Methods: The PTW inhomogeneity phantom (T40037) with different combination of heterogeneous slabs has been CT-scanned with Philips Brilliance 16 slice scanner. The phantom contains blocks of Rando Alderson materials mimicking lung, Polystyrene (Tissue), PTFE (Bone) and PMAA. The phantom has 30×30×2.5 cm base plate with 2cm recesses to insert inhomogeneity. The detector systems used in this study are diode, tlds and Gafchromic EBT2 films. The diode and tlds were included in CT scans. The CT sets are transferred to Eclipse treatment planningmore » system. Several plans have been created with Eclipse Monte Carlo (EMC) algorithm 11.0.21. Measurements have been carried out in Varian TrueBeam machine for energy from 6–22mev. Results: The measured and calculated doses agreed very well for tissue like media. The agreement was reasonably okay for the presence of lung inhomogeneity. The point dose agreement was within 3.5% and Gamma passing rate at 3%/3mm was greater than 93% except for 6Mev(85%). The disagreement can reach as high as 10% in the presence of bone inhomogeneity. This is due to eclipse reporting dose to the medium as opposed to the dose to the water as in conventional calculation engines. Conclusion: Care must be taken when using Varian Eclipse EMC algorithm for dose calculation for routine clinical uses. The algorithm dose not report dose to water in which most of the clinical experiences are based on rather it just reports dose to medium directly. In the presence of inhomogeneity such as bone, the dose discrepancy can be as high as 10% or even more depending on the location of normalization point or volume. As Radiation oncology as an empirical science, care must be taken before using EMC reported monitor units for clinical uses.« less

  8. SU-F-T-255: Accuracy and Precision of Dynamic Tracking Irradiation with VERO-4DRT System

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hayashi, N; Takada, Y; Mizuno, T

    2016-06-15

    Purpose: The VERO-4DRT system is able to provide dynamic tracking irradiation (DTI) for the target with respiratory motion. This technique requires enough commissioning for clinical implementation. The purpose of this study is to make sure the accuracy and precision of DTI using VERO- 4DRT through commissioning from fundamental evaluation to end-to-end test. Method: We evaluated several contents for DTI commissioning: the accuracy of absorption dose at isocenter in DTI, the field size and penumbra of DTI, the accuracy of 4D modeling in DTI. All evaluations were performed by respiratory motion phantom (Quasar phantom). These contents were compared the results betweenmore » static irradiation and DTI. The shape of radiation field was set to square from 3 cm × 3 cm to 10 cm × 10 cm. The micro 3D chamber and Gafchromic EBT3 film were used for absorbed dose and relative dose distribution measurement, respectively. The sine and irregular shaped waves were used for demonstrative respiratory motion. The visicoil was implanted into the phantom for guidance of respiratory motion. The respiration patterns of frequency and motion amount were set to 10–15 BPM and 1–2 cm, respectively. Results: As the result of absorbed dose of DTI in comparison with static irradiation, the average dose error at isocenter was 0.5% even though various respiratory patterns were set on. As the result of relative dose distribution, the field size (set it on 50% dose line) was not significantly changed in all respiratory patterns. However, the penumbra was larger in greater respiratory motion (up to 4.1 mm). The 4D modeling coincidence between actual and created waves was within 1%. Conclusion: The DTI using VERO-4DRT can provide sufficient accuracy and precision in absorbed dose and distribution. However, the patientspecific quantitative internal margin corresponding respiratory motion should be taken into consideration with image guidance.« less

  9. Grid therapy using high definition multileaf collimators: realizing benefits of the bystander effect.

    PubMed

    Peng, Valery; Suchowerska, Natalka; Rogers, Linda; Claridge Mackonis, Elizabeth; Oakes, Samantha; McKenzie, David R

    2017-08-01

    In microbeam radiotherapy (MRT), parallel arrays of high-intensity synchrotron x-ray beams achieve normal tissue sparing without compromising tumor control. Grid-therapy using clinical linacs has spatial modulation on a larger scale and achieves promising results for palliative treatments of bulky tumors. The availability of high definition multileaf collimators (HDMLCs) with 2.5 mm leaves provides an opportunity for grid-therapy to more closely approach MRT. However, challenges to the wider implementation of grid-therapy remain because spatial modulation of the target volume runs counter to current radiotherapy practice and mechanisms for the beneficial effects of MRT are not fully understood. Without more knowledge of cell dose responses, a quantitative basis for planning treatments is difficult. The aim of this study is to determine if therapeutic benefits of MRT can be achieved using a linac with HDMLCs and if so, to develop a predictive model to support treatment planning. HD120-MLCs of a Varian Novalis TX TM were used to generate grid patterns of 2.5 and 5.0 mm spacing, which were characterized dosimetrically using Gafchromic TM EBT3 film. Clonogenic survival of normal (HUVEC) and cancer (NCI-H460, HCC-1954) cell lines following irradiation under the grid and open fields using a 6 MV photon beam were compared in-vitro for the same average dose. Relative to an open field, survival of normal cells in a 2.5 mm striped field was the same, while the survival of both cancer cell lines was significantly lower. A mathematical model was developed to incorporate dose gradients of the spatial modulation into the standard linear quadratic model. Our new bystander extended LQ model assumes spatial gradients drive the diffusion of soluble factors that influence survival through bystander effects, successfully predicting the experimental results that show an increased therapeutic ratio. Our results challenge conventional radiotherapy practice and propose that

  10. An investigation of a PRESAGE® in-vivo dosimeter for brachytherapy

    PubMed Central

    Vidovic, A K; Juang, T; Meltsner, S; Adamovics, J; Chino, J; Steffey, B; Craciunescu, O; Oldham, M

    2014-01-01

    Determining accurate in-vivo dosimetry in brachytherapy treatment with high dose gradients is challenging. Here we introduce, investigate, and characterize a novel in-vivo dosimeter and readout technique with the potential to address this problem. A cylindrical (4 mm x 20 mm) tissue equivalent radiochromic dosimeter PRESAGE® In-Vivo (PRESAGE®-IV) is investigated. Two readout methods of the radiation induced change in optical density (OD) were investigated: (i) volume-averaged readout by spectrophotometer, and (ii) a line profile readout by 2D projection imaging utilizing a high-resolution (50 micron) telecentric optical system. Method (i) is considered the gold standard when applied to PRESAGE® in optical cuvettes. The feasibility of both methods was evaluated by comparison to standard measurements on PRESAGE® in optical cuvettes via spectrophotometer. An end-to-end feasibility study was performed by a side-by-side comparison with TLDs in an 192Ir HDR delivery. 7 and 8 Gy was delivered to PRESAGE®-IV and TLDs attached to the surface of a vaginal cylinder. Known geometry enabled direct comparison of measured dose with commissioned treatment planning system. A high-resolution readout study under a steep dose gradient region showed 98.9% (5%/1 mm) agreement between PRESAGE®-IV and Gafchromic® EBT2 Film. Spectrometer measurements exhibited a linear dose response between 0–15 Gy with sensitivity of 0.0133 ± 0.0007 ΔOD/(Gy·cm) at the 95% confidence interval. Method (ii) yielded a linear response with sensitivity of 0.0132 ± 0.0006 (ΔOD/Gy), within 2% of method (i). Method (i) has poor spatial resolution due to volume averaging. Method (ii) has higher resolution (~1mm) without loss of sensitivity or increased noise. Both readout methods are shown to be feasible. The end-to-end comparison revealed a 2.5% agreement between PRESAGE®-IV and treatment plan in regions of uniform high dose. PRESAGE®-IV shows promise for in-vivo dose verification, although

  11. SU-F-I-06: Evaluation of Imaging Dose for Modulation Layer Based Dual Energy Cone-Beam CT

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ju, Eunbin; Ahn, SoHyun; Cho, Samju

    Purpose: Dual energy cone beam CT system is finding a variety of promising applications in diagnostic CT, both in imaging of endogenous materials and exogenous materials across a range of body sites. Dual energy cone beam CT system to suggest in this study acquire image by rotating 360 degree with half of the X-ray window covered using copper modulation layer. In the region that covered by modulation layer absorb the low energy X-ray by modulation layer. Relative high energy X-ray passes through the layer and contributes to image reconstruction. Dose evaluation should be carried out in order to utilize suchmore » an imaging acquirement technology for clinical use. Methods: For evaluating imaging dose of modulation layer based dual energy cone beam CT system, Prototype cone beam CT that configured X-ray tube (D054SB, Toshiba, Japan) and detector (PaxScan 2520V, Varian Medical Systems, Palo Alto, CA) is used. A range of 0.5–2.0 mm thickness of modulation layer is implemented in Monte Carlo simulation (MCNPX, ver. 2.6.0, Los Alamos National Laboratory, USA) with half of X-ray window covered. In-house phantom using in this study that has 3 cylindrical phantoms configured water, Teflon air with PMMA covered for verifying the comparability the various material in human body and is implemented in Monte Carlo simulation. The actual dose with 2.0 mm copper covered half of X-ray window is measured using Gafchromic EBT3 film with 5.0 mm bolus for compared with simulative dose. Results: Dose in phantom reduced 33% by copper modulation layer of 2.0 mm. Scattering dose occurred in modulation layer by Compton scattering effect is 0.04% of overall dose. Conclusion: Modulation layer of that based dual energy cone beam CT has not influence on unnecessary scatter dose. This study was supported by the Radiation Safety Research Programs (1305033) through the Nuclear Safety and Security Commission.« less

  12. Comparison of measured and estimated maximum skin doses during CT fluoroscopy lung biopsies.

    PubMed

    Zanca, F; Jacobs, A; Crijns, W; De Wever, W

    2014-07-01

    To measure patient-specific maximum skin dose (MSD) associated with CT fluoroscopy (CTF) lung biopsies and to compare measured MSD with the MSD estimated from phantom measurements, as well as with the CTDIvol of patient examinations. Data from 50 patients with lung lesions who underwent a CT fluoroscopy-guided biopsy were collected. The CT protocol consisted of a low-kilovoltage (80 kV) protocol used in combination with an algorithm for dose reduction to the radiology staff during the interventional procedure, HandCare (HC). MSD was assessed during each intervention using EBT2 gafchromic films positioned on patient skin. Lesion size, position, total fluoroscopy time, and patient-effective diameter were registered for each patient. Dose rates were also estimated at the surface of a normal-size anthropomorphic thorax phantom using a 10 cm pencil ionization chamber placed at every 30°, for a full rotation, with and without HC. Measured MSD was compared with MSD values estimated from the phantom measurements and with the cumulative CTDIvol of the procedure. The median measured MSD was 141 mGy (range 38-410 mGy) while the median cumulative CTDIvol was 72 mGy (range 24-262 mGy). The ratio between the MSD estimated from phantom measurements and the measured MSD was 0.87 (range 0.12-4.1) on average. In 72% of cases the estimated MSD underestimated the measured MSD, while in 28% of the cases it overestimated it. The same trend was observed for the ratio of cumulative CTDIvol and measured MSD. No trend was observed as a function of patient size. On average, estimated MSD from dose rate measurements on phantom as well as from CTDIvol of patient examinations underestimates the measured value of MSD. This can be attributed to deviations of the patient's body habitus from the standard phantom size and to patient positioning in the gantry during the procedure.

  13. SU-F-T-211: Evaluation of a Dual Focusing Magnet System for the Treatment of Small Proton Targets

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Nguyen, TT; McAuley, GA; Heczko, S

    Purpose: To investigate magnetic focusing for small volume proton targets using a doublet combination of quadrupole rare earth permanent magnet Halbach cylinder assemblies Methods: Monte Carlo computer simulations were performed using the Geant4 toolkit to compare dose depositions of proton beams transported through two focusing magnets or in their absence. Proton beams with energies of 127 MeV and initial diameters of 5, 8 and 10 mm were delivered through two identical focusing magnets similar to those currently in experimental use at Loma Linda University Medical Center. Analogous experiments used optimized configurations based on the simulation results. Dose was measured bymore » a diode detector and Gafchromic EBT3 film and compared to simulation data. Based on results from the experimental data, an additional set of simulations was performed with an initial beam diameter of 18 mm and a two differing length magnets (40mm & 68mm). Results: Experimental data matched well with Monte Carlo simulations. However, under conditions necessary to produce circular beam spots at target depth, magnetically focused beams using two identical 40 mm length magnets did not meet all of our performance criteria of circular beam spots, improved peak to entrance (P/E) dose ratios and dose delivery efficiencies. The simulations using the longer 68 mm 2nd magnet yielded better results with 34% better P/E dose ratio and 20–50% better dose delivery efficiencies when compared to unfocused 10 mm beams. Conclusion: While magnetic focusing using two magnets with identical focusing power did not yield desired results, ongoing Monte Carlo simulations suggest that increasing the length of the 2nd magnet to 68 mm could improve P/E dose ratios and dose efficiencies. Future work includes additional experimental validation of the longer 2nd magnet setup as well as experiments with triplet magnet systems. This project was sponsored with funding from the Department of Defense (DOD# W81XWH-BAA-10

  14. An investigation of a PRESAGE® in vivo dosimeter for brachytherapy

    NASA Astrophysics Data System (ADS)

    Vidovic, A. K.; Juang, T.; Meltsner, S.; Adamovics, J.; Chino, J.; Steffey, B.; Craciunescu, O.; Oldham, M.

    2014-07-01

    Determining accurate in vivo dosimetry in brachytherapy treatment with high dose gradients is challenging. Here we introduce, investigate, and characterize a novel in vivo dosimeter and readout technique with the potential to address this problem. A cylindrical (4 mm × 20 mm) tissue equivalent radiochromic dosimeter PRESAGE® in vivo (PRESAGE®-IV) is investigated. Two readout methods of the radiation induced change in optical density (OD) were investigated: (i) volume-averaged readout by spectrophotometer, and (ii) a line profile readout by 2D projection imaging utilizing a high-resolution (50 micron) telecentric optical system. Method (i) is considered the gold standard when applied to PRESAGE® in optical cuvettes. The feasibility of both methods was evaluated by comparison to standard measurements on PRESAGE® in optical cuvettes via spectrophotometer. An end-to-end feasibility study was performed by a side-by-side comparison with TLDs in an 192Ir HDR delivery. 7 and 8 Gy was delivered to PRESAGE®-IV and TLDs attached to the surface of a vaginal cylinder. Known geometry enabled direct comparison of measured dose with a commissioned treatment planning system. A high-resolution readout study under a steep dose gradient region showed 98.9% (5%/1 mm) agreement between PRESAGE®-IV and Gafchromic® EBT2 Film. Spectrometer measurements exhibited a linear dose response between 0-15 Gy with sensitivity of 0.0133 ± 0.0007 ΔOD/(Gy ṡ cm) at the 95% confidence interval. Method (ii) yielded a linear response with sensitivity of 0.0132 ± 0.0006 (ΔOD/Gy), within 2% of method (i). Method (i) has poor spatial resolution due to volume averaging. Method (ii) has higher resolution (˜1 mm) without loss of sensitivity or increased noise. Both readout methods are shown to be feasible. The end-to-end comparison revealed a 2.5% agreement between PRESAGE®-IV and treatment plan in regions of uniform high dose. PRESAGE®-IV shows promise for in vivo dose verification

  15. Intensity modulated operating mode of the rotating gamma system.

    PubMed

    Sengupta, Bishwambhar; Gulyas, Laszlo; Medlin, Donald; Koroknai, Tibor; Takacs, David; Filep, Gyorgy; Panko, Peter; Godo, Bence; Hollo, Tamas; Zheng, Xiao Ran; Fedorcsak, Imre; Dobai, Jozsef; Bognar, Laszlo; Takacs, Endre

    2018-05-01

    The purpose of this work was to explore two novel operation modalities of the rotating gamma systems (RGS) that could expand its clinical application to lesions in close proximity to critical organs at risk (OAR). The approach taken in this study consists of two components. First, a Geant4-based Monte Carlo (MC) simulation toolkit is used to model the dosimetric properties of the RGS Vertex 360™ for the normal, intensity modulated radiosurgery (IMRS), and speed modulated radiosurgery (SMRS) operation modalities. Second, the RGS Vertex 360™ at the Rotating Gamma Institute in Debrecen, Hungary is used to collect experimental data for the normal and IMRS operation modes. An ion chamber is used to record measurements of the absolute dose. The dose profiles are measured using Gafchromic EBT3 films positioned within a spherical water equivalent phantom. A strong dosimetric agreement between the measured and simulated dose profiles and penumbra was found for both the normal and IMRS operation modes for all collimator sizes (4, 8, 14, and 18 mm diameter). The simulated falloff and maximum dose regions agree better with the experimental results for the 4 and 8 mm diameter collimators. Although the falloff regions align well in the 14 and 18 mm collimators, the maximum dose regions have a larger difference. For the IMRS operation mode, the simulated and experimental dose distributions are ellipsoidal, where the short axis aligns with the blocked angles. Similarly, the simulated dose distributions for the SMRS operation mode also adopt an ellipsoidal shape, where the short axis aligns with the angles where the orbital speed is highest. For both modalities, the dose distribution is highly constrained with a sharper penumbra along the short axes. Dose modulation of the RGS can be achieved with the IMRS and SMRS modes. By providing a highly constrained dose distribution with a sharp penumbra, both modes could be clinically applicable for the treatment of lesions in close

  16. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kakinohana, Y; Toita, T; Kasuya, G

    Purpose: To compare the dosimetric properties of radiochromic films with different orientation. Methods: A sheet of EBT3 film was cut into eight pieces with the following sizes: 15×15 cm2 (one piece), 5x15 cm{sup 2} (two) and 4×5 cm{sup 2} (five). A set of two EBT3 sheets was used at each dose level. Two sets were used changing the delivered doses (1 and 2 Gy). The 5×15 cm{sup 2} pieces were rotated by 90 degrees in relation to each other, such that one had landscape orientation and the other had portrait orientation. All 5×15 cm2 pieces were irradiated with their longmore » side aligned with the x-axis of the radiation field. The 15×15 cm{sup 2} pieces were irradiated rotated at 90 degrees to each other. Five pieces, (a total of ten from two sheets) were used to obtain a calibration curve. The irradiated films were scanned using an Epson ES-2200 scanner and were analyzed using ImageJ software. In this study, no correction was applied for the nonuniform scanner signal that is evident in the direction of the scanner lamp. Each film piece was scanned both in portrait and landscape orientations. Dosimetric comparisons of the beam profiles were made in terms of the film orientations (portrait and landscape) and scanner bed directions (perpendicular and parallel to the scanner movement). Results: In general, portrait orientation exhibited higher noise than landscape and was adversely affected to a great extent by the nonuniformity in the direction of the scanner lamp. A significant difference in the measured field widths between the perpendicular and parallel directions was found for both orientations. Conclusion: Without correction for the nonuniform scanner signal in the direction of the scanner lamp, a landscape orientation is preferable. A more detailed investigation is planned to evaluate quantitatively the effect of orientation on the dosimetric properties of a film.« less

  17. Surface-enhanced Raman scattering and DFT investigation of Eriochrome Black T metal chelating compound

    NASA Astrophysics Data System (ADS)

    Szabó, László; Herman, Krisztian; Leopold, Nicolae; Buzumurgă, Claudia; Chiş, Vasile

    2011-06-01

    The surface-enhanced Raman scattering (SERS) spectra of Eriochrome Black T (EBT) and its Cu(II), Fe(III), Mn(II) and Pb(II) complexes were recorded using a hydroxylamine reduced silver colloid. Molecular geometry optimization, molecular electrostatic potential (MEP) distribution and vibrational frequencies calculation were performed at B3LYP/6-31G(d) level of theory for the EBT molecule and its Cu(EBT), Fe(EBT) and Mn(EBT) metal complexes. Differentiation between EBT complexes of Cu(II), Fe(III), Mn(II) and Pb(II) is shown by the SERS spectral features of each complex.

  18. Innovative methodology for intercomparison of radionuclide calibrators using short half-life in situ prepared radioactive sources

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Oliveira, P. A.; Santos, J. A. M., E-mail: joao.santos@ipoporto.min-saude.pt; Serviço de Física Médica do Instituto Português de Oncologia do Porto Francisco Gentil, EPE, Porto

    2014-07-15

    Purpose: An original radionuclide calibrator method for activity determination is presented. The method could be used for intercomparison surveys for short half-life radioactive sources used in Nuclear Medicine, such as{sup 99m}Tc or most positron emission tomography radiopharmaceuticals. Methods: By evaluation of the resulting net optical density (netOD) using a standardized scanning method of irradiated Gafchromic XRQA2 film, a comparison of the netOD measurement with a previously determined calibration curve can be made and the difference between the tested radionuclide calibrator and a radionuclide calibrator used as reference device can be calculated. To estimate the total expected measurement uncertainties, a carefulmore » analysis of the methodology, for the case of{sup 99m}Tc, was performed: reproducibility determination, scanning conditions, and possible fadeout effects. Since every factor of the activity measurement procedure can influence the final result, the method also evaluates correct syringe positioning inside the radionuclide calibrator. Results: As an alternative to using a calibrated source sent to the surveyed site, which requires a relatively long half-life of the nuclide, or sending a portable calibrated radionuclide calibrator, the proposed method uses a source preparedin situ. An indirect activity determination is achieved by the irradiation of a radiochromic film using {sup 99m}Tc under strictly controlled conditions, and cumulated activity calculation from the initial activity and total irradiation time. The irradiated Gafchromic film and the irradiator, without the source, can then be sent to a National Metrology Institute for evaluation of the results. Conclusions: The methodology described in this paper showed to have a good potential for accurate (3%) radionuclide calibrators intercomparison studies for{sup 99m}Tc between Nuclear Medicine centers without source transfer and can easily be adapted to other short half

  19. Black Films and Film-Makers.

    ERIC Educational Resources Information Center

    Patterson, Lindsay, Ed.

    The development of black films and the attitudes of the film industry toward black films and black actors are some of the topics examined in this anthology of essays. Section 1, "Nigger to Supernigger," contains such articles as "The Death of Rastus: Negroes in American Films" by Thomas R. Cripps and "Folk Values in a New Medium" by Alain Locke…

  20. SU-E-T-361: Energy Dependent Radiation/light-Field Misalignment On Truebeam Linear Accelerator

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sperling, N; Tanny, S; Parsai, E

    2015-06-15

    Purpose: Verifying the co-incidence of the radiation and light field is recommended by TG-142 for monthly and annual checks. On a digital accelerator, it is simple to verify that beam steering settings are consistent with accepted and commissioned values. This fact should allow for physicists to verify radiation-light-field co-incidence for a single energy and accept that Result for all energies. We present a case where the radiation isocenter deviated for a single energy without any apparent modification to the beam steering parameters. Methods: The radiation isocenter was determined using multiple Methods: Gafchromic film, a BB test, and radiation profiles measuredmore » with a diode. Light-field borders were marked on Gafchromic film and then irradiated for all photon energies. Images of acceptance films were compared with films taken four months later. A phantom with a radio-opaque BB was aligned to isocenter using the light-field and imaged using the EPID for all photon energies. An unshielded diode was aligned using the crosshairs and then beam profiles of multiple field sizes were obtained. Field centers were determined using Omni-Pro v7.4 software, and compared to similar scans taken during commissioning. Beam steering parameter files were checked against backups to determine that the steering parameters were unchanged. Results: There were no differences between the configuration files from acceptance. All three tests demonstrated that a single energy had deviated from accepted values by 0.8 mm in the inline direction. The other two energies remained consistent with previous measurements. The deviated energy was re-steered to be within our clinical tolerance. Conclusions: Our study demonstrates that radiation-light-field coincidence is an energy dependent effect for modern linacs. We recommend that radiation-light-field coincidence be verified for all energies on a monthly basis, particularly for modes used to treat small fields, as these may drift

  1. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Thomas, S; Yuen, C; Huang, V

    Purpose: In this abstract we implement and validate a 4D VMAT Acuros XB dose calculation using Gafchromic film. Special attention is paid to the physical material assignment in the CT dataset and to reported dose to water and dose to medium. Methods: A QUASAR phantom with a 3 cm sinusoidal tumor motion and 5 second period was scanned using 4D computed tomography. A CT was also obtained of the static QUASAR phantom with the tumor at the central position. A VMAT plan was created on the average CT dataset and was delivered on a Varian TrueBeam linear accelerator. The trajectorymore » log file from this treatment was acquired and used to create 10 VMAT subplans (one for each portion of the breathing cycle). Motion for each subplan was simulated by moving the beam isocentre in the superior/inferior direction in the Treatment Planning System on the static CT scan. The 10 plans were calculated (both dose to medium and dose to water) and summed for 1) the original HU values from the static CT scan and 2) the correct physical material assignment in the CT dataset. To acquire a breathing phase synchronized film measurements the trajectory log was used to create a VMAT delivery plan which includes dynamic couch motion using the Developer Mode. Three different treatment start phases were investigated (mid inhalation, full inhalation and full exhalation). Results: For each scenario the coronal dose distributions were measured using Gafchromic film and compared to the corresponding calculation with Film QA Pro Software using a Gamma test with a 3%/3mm distance to agreement criteria. Good agreement was found between calculation and measurement. No statistically significant difference in agreement was found between calculations to original HU values vs calculations to over-written (material-assigned) HU values. Conclusion: The investigated 4D dose calculation method agrees well with measurement.« less

  2. WE-F-16A-01: Commissioning and Clinical Use of PC-ISO for Customized, 3D Printed, Gynecological Brachytherapy Applicators

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Cunha, J; Sethi, R; Mellis, K

    Purpose: (1) Evaluate the safety and radiation attenuation properties of PCISO, a bio-compatible, sterilizable 3D printing material by Stratasys, (2) establish a method for commissioning customized multi- and single-use 3D printed applicators, (3) report on use of customized vaginal cylinders used to treat a series of serous endometrial cancer patient. Methods: A custom film dosimetry apparatus was designed to hold a Gafchromic radio film segment between two blocks of PC-ISO and 3D-printed using a Fortus 400mc (StrataSys). A dose plan was computed using 13 dwell positions at 2.5 mm spacing and normalized to 1500 cGy at 1 cm. Film exposuremore » was compared to control tests in only air and only water. The average Hounsfield Unit (HU) was computed and used to verify water equivalency. For the clinical use cases, the physician specifies the dimensions and geometry of a custom applicator from which a CAD model is designed and printed. Results: The doses measured from the PC-ISO Gafchromic film test were within 1% of the dose measured in only water between 1cm and 6cm from the channel. Doses increased 7–4% measured in only air. HU range was 11–43. The applicators were sterilized using the Sterrad system multiple times without damage. As of submission 3 unique cylinders have been designed, printed, and used in the clinic. A standardizable workflow for commissioning custom 3D printed applicators was codified and will be reported. Conclusions: Quality assurance (QA) evaluation of the PC-ISO 3D-printing material showed that PC-ISO is a suitable material for a gynecological brachytherapy vaginal cylinder in a clinical setting. With the material commissioning completed, if the physician determines that a better treatment would Result, a customized design is fabricated with limited additional QA necessary. Although this study was specific to PC-ISO, the same setup can be used to evaluate other 3D-printing materials.« less

  3. TH-EF-BRB-10: Dosimetric Validation of a Trajectory Based Cranial SRS Treatment Technique On a Varian TrueBeam Linac

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wilson, B; Vancouver Cancer Centre, Vancouver, BC; Gete, E

    2016-06-15

    Purpose: This work investigates the dosimetric accuracy of a trajectory based delivery technique in which an optimized radiation beam is delivered along a Couch-Gantry trajectory that is formed by simultaneous rotation of the linac gantry and the treatment couch. Methods: Nine trajectory based cranial SRS treatment plans were created using in-house optimization software. The plans were calculated for delivery on the TrueBeam STx linac with 6MV photon beam. Dose optimization was performed along a user-defined trajectory using MLC modulation, dose rate modulation and jaw tracking. The pre-defined trajectory chosen for this study is formed by a couch rotation through itsmore » full range of 180 degrees while the gantry makes four partial arc sweeps which are 170 degrees each. For final dose calculation, the trajectory based plans were exported to the Varian Eclipse Treatment Planning System. The plans were calculated on a homogeneous cube phantom measuring 18.2×18.2×18.2 cm3 with the analytical anisotropic algorithm (AAA) using a 1mm3 calculation voxel. The plans were delivered on the TrueBeam linac via the developer’s mode. Point dose measurements were performed on 9 patients with the IBA CC01 mini-chamber with a sensitive volume of 0.01 cc. Gafchromic film measurements along the sagittal and coronal planes were performed on three of the 9 treatment plans. Point dose values were compared with ion chamber measurements. Gamma analysis comparing film measurement and AAA calculations was performed using FilmQA Pro. Results: The AAA calculations and measurements were in good agreement. The point dose difference between AAA and ion chamber measurements were within 2.2%. Gamma analysis test pass rates (2%, 2mm passing criteria) for the Gafchromic film measurements were >95%. Conclusion: We have successfully tested TrueBeam’s ability to deliver accurate trajectory based treatments involving simultaneous gantry and couch rotation with MLC and dose rate modulation

  4. SU-E-T-651: Quantification of Dosimetric Accuracy of Respiratory Gated Stereotactic Body Radiation Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Thiyagarajan, Rajesh; Vikraman, S; Maragathaveni, S

    2015-06-15

    Purpose: To quantify the dosimetric accuracy of respiratory gated stereotactic body radiation therapy delivery using dynamic thorax phantom. Methods: Three patients with mobile target (2 lung, 1liver) were chosen. Retrospective 4DCT image sets were acquired for using Varian RPM system. An in-house MATLAB program was designed for MIP, MinIP and AvgIP generation. ITV was contoured on MIP image set for lung patients and on MinIP for liver patient. Dynamic IMRT plans were generated on selected phase bin image set in Eclipse (v10.0) planning system. CIRS dynamic thorax phantom was used to perform the dosimetric quality assurance. Patient breathing pattern filemore » from RPM system was converted to phantom compatible file by an in-house MATLAB program. This respiratory pattern fed to the CIRS dynamic thorax phantom. 4DCT image set was acquired for this phantom using patient breathing pattern. Verification plans were generated using patient gating window and delivered on the phantom. Measurements were carried out using with ion chamber and EBT2 film. Exposed films were analyzed and evaluated in FilmQA software. Results: The stability of gated output in comparison with un-gated output was within 0.5%. The Ion chamber measured and TPS calculated dose compared for all the patients. The difference observed was 0.45%, −0.52% and −0.54 for Patient 1, Patient2 and Patient 3 respectively.Gamma value evaluated from EBT film shows pass rates from 92.41% to 99.93% for 3% dose difference and 3mm distance to agreement criteria. Conclusion: Dosimetric accuracy of respiratory gated SBRT delivery for lung and liver was dosimetrically acceptable. The Ion chamber measured dose was within 0.203±0.5659% of the expected dose. Gamma pass rates were within 96.63±3.84% of the expected dose.« less

  5. Dosimetric evaluation of internal shielding in a high dose rate skin applicator

    PubMed Central

    Granero, Domingo; Perez-Calatayud, Jose; Carmona, Vicente; Pujades, M Carmen; Ballester, Facundo

    2011-01-01

    Purpose The Valencia HDR applicators are accessories of the microSelectron HDR afterloading system (Nucletron) shaped as truncated cones. The base of the cone is either 2 or 3 cm diameter. They are intended to treat skin lesions, being the typical prescription depth 3 mm. In patients with eyelid lesions, an internal shielding is very useful to reduce the dose to the ocular globe. The purpose of this work was to evaluate the dose enhancement from potential backscatter and electron contamination due to the shielding. Material and methods Two methods were used: a) Monte Carlo simulation, performed with the GEANT4 code, 2 cm Valencia applicator was placed on the surface of a water phantom in which 2 mm lead slab was located at 3 mm depth; b) radiochromic EBT films, used to verify the Monte Carlo results, positioning the films at 1.5, 3, 5 and 7 mm depth, inside the phantom. Two irradiations, with and without the lead shielding slab, were carried out. Results The Monte Carlo results showed that due to the backscatter component from the lead, the dose level raised to about 200% with a depth range of 0.5 mm. Under the lead the dose level was enhanced to about 130% with a depth range of 1 mm. Two millimeters of lead reduce the dose under the slab with about 60%. These results agree with film measurements within uncertainties. Conclusions In conclusion, the use of 2 mm internal lead shielding in eyelid skin treatments with the Valencia applicators were evaluated using MC methods and EBT film dosimetry. The minimum bolus thickness that was needed above and below the shielding was 0.5 mm and 1 mm respectively, and the shielding reduced the absorbed dose delivered to the ocular globe by about 60%. PMID:27877198

  6. SU-F-SPS-11: The Dosimetric Comparison of Truebeam 2.0 and Cyberknife M6 Treatment Plans for Brain SRS Treatment

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Mabhouti, H; Sanli, E; Cebe, M

    Purpose: Brain stereotactic radiosurgery involves the use of precisely directed, single session radiation to create a desired radiobiologic response within the brain target with acceptable minimal effects on surrounding structures or tissues. In this study, the dosimetric comparison of Truebeam 2.0 and Cyberknife M6 treatment plans were made. Methods: For Truebeam 2.0 machine, treatment planning were done using 2 full arc VMAT technique with 6 FFF beam on the CT scan of Randophantom simulating the treatment of sterotactic treatments for one brain metastasis. The dose distribution were calculated using Eclipse treatment planning system with Acuros XB algorithm. The treatment planningmore » of the same target were also done for Cyberknife M6 machine with Multiplan treatment planning system using Monte Carlo algorithm. Using the same film batch, the net OD to dose calibration curve was obtained using both machine by delivering 0- 800 cGy. Films were scanned 48 hours after irradiation using an Epson 1000XL flatbed scanner. Dose distribution were measured using EBT3 film dosimeter. The measured and calculated doses were compared. Results: The dose distribution in the target and 2 cm beyond the target edge were calculated on TPSs and measured using EBT3 film. For cyberknife plans, the gamma analysis passing rates between measured and calculated dose distributions were 99.2% and 96.7% for target and peripheral region of target respectively. For Truebeam plans, the gamma analysis passing rates were 99.1% and 95.5% for target and peripheral region of target respectively. Conclusion: Although, target dose distribution calculated accurately by Acuros XB and Monte Carlo algorithms, Monte carlo calculation algorithm predicts dose distribution around the peripheral region of target more accurately than Acuros algorithm.« less

  7. Clinicopathologic Comparison of High-Dose-Rate Endorectal Brachytherapy versus Conventional Chemoradiotherapy in the Neoadjuvant Setting for Resectable Stages II and III Low Rectal Cancer

    PubMed Central

    Smith, Jessica A.; Wild, Aaron T.; Singhi, Aatur; Raman, Siva P.; Qiu, Haoming; Kumar, Rachit; Hacker-Prietz, Amy; Hruban, Ralph H.; Kamel, Ihab R.; Efron, Jonathan; Wick, Elizabeth C.; Azad, Nilofer S.; Diaz, Luis A.; Le, Yi; Armour, Elwood P.; Gearhart, Susan L.; Herman, Joseph M.

    2012-01-01

    Purpose. To assess for differences in clinical, radiologic, and pathologic outcomes between patients with stage II-III rectal adenocarcinoma treated neoadjuvantly with conventional external beam radiotherapy (3D conformal radiotherapy (3DRT) or intensity-modulated radiotherapy (IMRT)) versus high-dose-rate endorectal brachytherapy (EBT). Methods. Patients undergoing neoadjuvant EBT received 4 consecutive daily 6.5 Gy fractions without chemotherapy, while those undergoing 3DRT or IMRT received 28 daily 1.8 Gy fractions with concurrent 5-fluorouracil. Data was collected prospectively for 7 EBT patients and retrospectively for 25 historical 3DRT/IMRT controls. Results. Time to surgery was less for EBT compared to 3DRT and IMRT (P < 0.001). There was a trend towards higher rate of pathologic CR for EBT (P = 0.06). Rates of margin and lymph node positivity at resection were similar for all groups. Acute toxicity was less for EBT compared to 3DRT and IMRT (P = 0.025). Overall and progression-free survival were noninferior for EBT. On MRI, EBT achieved similar complete response rate and reduction in tumor volume as 3DRT and IMRT. Histopathologic comparison showed that EBT resulted in more localized treatment effects and fewer serosal adhesions. Conclusions. EBT offers several practical benefits over conventional radiotherapy techniques and appears to be at least as effective against low rectal cancer as measured by short-term outcomes. PMID:22830003

  8. Feasibility of using Geant4 Monte Carlo simulation for IMRT dose calculations for the Novalis Tx with a HD-120 multi-leaf collimator

    NASA Astrophysics Data System (ADS)

    Jung, Hyunuk; Shin, Jungsuk; Chung, Kwangzoo; Han, Youngyih; Kim, Jinsung; Choi, Doo Ho

    2015-05-01

    The aim of this study was to develop an independent dose verification system by using a Monte Carlo (MC) calculation method for intensity modulated radiation therapy (IMRT) conducted by using a Varian Novalis Tx (Varian Medical Systems, Palo Alto, CA, USA) equipped with a highdefinition multi-leaf collimator (HD-120 MLC). The Geant4 framework was used to implement a dose calculation system that accurately predicted the delivered dose. For this purpose, the Novalis Tx Linac head was modeled according to the specifications acquired from the manufacturer. Subsequently, MC simulations were performed by varying the mean energy, energy spread, and electron spot radius to determine optimum values of irradiation with 6-MV X-ray beams by using the Novalis Tx system. Computed percentage depth dose curves (PDDs) and lateral profiles were compared to the measurements obtained by using an ionization chamber (CC13). To validate the IMRT simulation by using the MC model we developed, we calculated a simple IMRT field and compared the result with the EBT3 film measurements in a water-equivalent solid phantom. Clinical cases, such as prostate cancer treatment plans, were then selected, and MC simulations were performed. The accuracy of the simulation was assessed against the EBT3 film measurements by using a gamma-index criterion. The optimal MC model parameters to specify the beam characteristics were a 6.8-MeV mean energy, a 0.5-MeV energy spread, and a 3-mm electron radius. The accuracy of these parameters was determined by comparison of MC simulations with measurements. The PDDs and the lateral profiles of the MC simulation deviated from the measurements by 1% and 2%, respectively, on average. The computed simple MLC fields agreed with the EBT3 measurements with a 95% passing rate with 3%/3-mm gamma-index criterion. Additionally, in applying our model to clinical IMRT plans, we found that the MC calculations and the EBT3 measurements agreed well with a passing rate of greater

  9. Investigation of intracranial peripheral dose arising from the treatment of large lesions with Leksell GammaKnife Perfexion.

    PubMed

    Ruschin, Mark; Nordström, Håkan; Kjäll, Per; Cho, Young-Bin; Jaffray, David

    2009-06-01

    This investigation involves quantifying the extent of intracranial peripheral dose arising from simulated targets situated in the skull-base or upper-spine region using the Leksell GammaKnife Perfexion treatment unit. For each of three spherical target volumes--denoted as Vs (4 cm3), VM (18 cm3), and VL (60 cm3)--three treatment plans were manually generated, one for each of the three collimator sizes--4, 8, and 16 mm. Each of the plans was delivered to a spherical dosimetry phantom with an insert containing EBT Gafchromic film. The total dose at 70 mm from the targets' edges, %D(70 mm), was measured as a function of elevation angle and expressed as a percentage of the prescription dose. The film insert was placed centered in the median sagittal plane (Leksell X = 100) and %D(70 mm) was measured for the angular range from 0 degree (superior/along Z axis) to 90 degrees (anterior/along Y axis). For a given collimator i, the irradiation time ti to treat a spherical target of volume V using the 50% isodose line was observed to follow a power-law relationship of the form ti = Ai(V/ Vi)n where Ai was the maximum dose divided by collimator dose rate and Vi was the volume encompassed by the 50% isodose line for a single shot. The mean value of n was 0.61 (range: 0.61-0.62). Along the superior (Z) direction (angle=0 degree) and up to angles of around 30 degrees, the %D(70 mm) was always highest for the 4 mm plans, followed by the 8 mm, followed by the 16 mm. In this angular range, the maximum measured %D(70 mm) was 1.7% of the prescription dose. The intracranial peripheral dose along the superior direction (combined scatter and leakage dose) resulting from irradiation of upper-spine or base-of-skull lesions is measured to be less than 2% of the prescription dose, even for very large (60 cm3) targets. The results of this study indicate that, for a given target volume, treatment plans consisting of only 4 mm shots yield larger peripheral dose in the superior direction than 8

  10. Involving Employers in Training: Best Practices. Research and Evaluation Report Series 97-I.

    ERIC Educational Resources Information Center

    Isbell, Kellie; Trutko, John W.; Barnow, Burt S.; Nightengale, Demetra; Pindus, Nancy

    A review of the literature on employer-based training (EBT) and case studies of 17 exemplary EBT programs (9 of which were Job Training Partnership Act [JTPA] programs) were used to determine best practices in EBT and make recommendations for effective strategies for implementing EBT programs. Some of these strategies included the following: (1)…

  11. Dosimetric changes with computed tomography automatic tube-current modulation techniques.

    PubMed

    Spampinato, Sofia; Gueli, Anna Maria; Milone, Pietro; Raffaele, Luigi Angelo

    2018-04-06

    The study is aimed at a verification of dose changes for a computed tomography automatic tube-current modulation (ATCM) technique. For this purpose, anthropomorphic phantom and Gafchromic ® XR-QA2 films were used. Radiochromic films were cut according to the shape of two thorax regions. The ATCM algorithm is based on noise index (NI) and three exam protocols with different NI were chosen, of which one was a reference. Results were compared with dose values displayed by the console and with Poisson statistics. The information obtained with radiochromic films has been normalized with respect to the NI reference value to compare dose percentage variations. Results showed that, on average, the information reported by the CT console and calculated values coincide with measurements. The study allowed verification of the dose information reported by the CT console for an ATCM technique. Although this evaluation represents an estimate, the method can be a starting point for further studies.

  12. Technical note: patient-specific quality assurance methods for TomoDirect(TM) whole breast treatment delivery.

    PubMed

    Catuzzo, P; Zenone, F; Aimonetto, S; Peruzzo, A; Casanova Borca, V; Pasquino, M; Franco, P; La Porta, M R; Ricardi, U; Tofani, S

    2012-07-01

    To investigate the feasibility of implementing a novel approach for patient-specific QA of TomoDirect(TM) whole breast treatment. The most currently used TomoTherapy DQA method, consisting in the verification of the 2D dose distribution in a coronal or sagittal plane of the Cheese Phantom by means of gafchromic films, was compared with an alternative approach based on the use of two commercially available diode arrays, MapCHECK2(TM) and ArcCHECK(TM). The TomoDirect(TM) plans of twenty patients with a primary unilateral breast cancer were applied to a CT scan of the Cheese Phantom and a MVCT dataset of the diode arrays. Then measurements of 2D dose distribution were performed and compared with the calculated ones using the gamma analysis method with different sets of DTA and DD criteria (3%-3 mm, 3%-2 mm). The sensitivity of the diode arrays to detect delivery and setup errors was also investigated. The measured dose distributions showed excellent agreement with the TPS calculations for each detector, with averaged fractions of passed Γ values greater than 95%. The percentage of points satisfying the constraint Γ < 1 was significantly higher for MapCHECK2(TM) than for ArcCHECK(TM) and gafchromic films using both the 3%-3 mm and 3%-2 mm gamma criteria. Both the diode arrays show a good sensitivity to delivery and setup errors using a 3%-2 mm gamma criteria. MapCHECK2™ and ArcCHECK(TM) may fulfill the demands of an adequate system for TomoDirect(TM) patient-specific QA.

  13. Poster — Thur Eve — 28: Enabling trajectory-based radiotherapy on a TrueBeam accelerator with the Eclipse treatment planning system

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Mullins, J; Asiev, K; DeBlois, F

    2014-08-15

    The TrueBeam linear accelerator platform has a developer's mode which permits the user dynamic control over many of the machine's mechanical and radiation systems. Using this research tool, synchronous couch and gantry motion can be programmed to simulate isocentric treatment with a shortened SAD, with benefits such as smaller projected MLC leaf widths and an increased dose rate. In this work, water tank measurements were used to commission a virtual linear accelerator with an 85 cm SAD in Eclipse, from which several arc-based radiotherapy treatments were generated, including an inverse optimized VMAT delivery. For each plan, the pertinent treatment deliverymore » information was extracted from control points specified in the Eclipse-exported DICOM files using the pydicom package in Python, allowing construction of an XML control file. The dimensions of the jaws and MLC positions, defined for an 85 cm SAD in Eclipse, were scaled for delivery on a conventional SAD linear accelerator, and translational couch motion was added as a function of gantry angle to simulate delivery at 85 cm SAD. Ionization chamber and Gafchromic film measurements were used to compare the radiation delivery to dose calculations in Eclipse. With the exception of the VMAT delivery, ionization chamber measurements agreed within 3.3% of the Eclipse calculations. For the VMAT delivery, the ionization chamber was located in an inhomogeneous region, but gamma evaluation of the Gafchromic film plane resulted in a 94.5% passing rate using criteria of 3 mm/3%. The results indicate that Eclipse calculation infrastructure can be used.« less

  14. Availability of a containerless polymer gel detector and a gelatin container

    NASA Astrophysics Data System (ADS)

    Tominaga, Takahiro; Yoshioka, Munenori; Hayashi, Shin-ichiro; Usui, Shuji; Tada, Mitsutoshi

    2015-01-01

    We considered an availability of the polymer gel detector without container but with a plastic wrap under assumption of the low oxygen transmissivity of a sheet of plastic wrap. And a gelatin container was also examined for a gel detector. These samples can be made easily and this containerless polymer gel detector well works without any artifacts by means of wrapping with a thin plastic sheet. Nevertheless, there is still room for improvement on preventing oxygen contamination. Combination with a gelatin container and a polymer gel detector and/or Gafchromic films has a various potential for extension of 3D dosimetry.

  15. Influence of film dimensions on film droplet formation.

    PubMed

    Holmgren, Helene; Ljungström, Evert

    2012-02-01

    Aerosol particles may be generated from rupturing liquid films through a droplet formation mechanism. The present work was undertaken with the aim to throw some light on the influence of film dimensions on droplet formation with possible consequences for exhaled breath aerosol formation. The film droplet formation process was mimicked by using a purpose-built device, where fluid films were spanned across holes of known diameters. As the films burst, droplets were formed and the number and size distributions of the resulting droplets were determined. No general relation could be found between hole diameter and the number of droplets generated per unit surface area of fluid film. Averaged over all film sizes, a higher surface tension yielded higher concentrations of droplets. Surface tension did not influence the resulting droplet diameter, but it was found that smaller films generated smaller droplets. This study shows that small fluid films generate droplets as efficiently as large films, and that droplets may well be generated from films with diameters below 1 mm. This has implications for the formation of film droplets from reopening of closed airways because human terminal bronchioles are of similar dimensions. Thus, the results provide support for the earlier proposed mechanism where reopening of closed airways is one origin of exhaled particles.

  16. National survey of psychotherapy training in psychiatry, psychology, and social work.

    PubMed

    Weissman, Myrna M; Verdeli, Helen; Gameroff, Marc J; Bledsoe, Sarah E; Betts, Kathryn; Mufson, Laura; Fitterling, Heidi; Wickramaratne, Priya

    2006-08-01

    Approximately 3% of the US population receives psychotherapy each year from psychiatrists, psychologists, or social workers. A modest number of psychotherapies are evidence-based therapy (EBT) in that they have been defined in manuals and found efficacious in at least 2 controlled clinical trials with random assignment that include a control condition of psychotherapy, placebo, pill, or other treatment and samples of sufficient power with well-characterized patients. Few practitioners use EBT. To determine the amount of EBT taught in accredited training programs in psychiatry, psychology (PhD and PsyD), and social work and to note whether the training was elective or required and presented as a didactic (coursework) or clinical supervision. A cross-sectional survey of a probability sample of all accredited training programs in psychiatry, psychology, and social work in the United States. Responders included training directors (or their designates) from 221 programs (73 in psychiatry, 63 in PhD clinical psychology, 21 in PsyD psychology, and 64 in master's-level social work). The overall response rate was 73.7%. Main Outcome Measure Requiring both a didactic and clinical supervision in an EBT. Although programs offered electives in EBT and non-EBT, few required both a didactic and clinical supervision in EBT, and most required training was non-EBT. Psychiatry required coursework and clinical supervision in the largest percentage of EBT (28.1%). Cognitive behavioral therapy was the EBT most frequently offered and required as a didactic in all 3 disciplines. More than 90% of the psychiatry training programs were complying with the new cognitive behavior therapy requirement. The 2 disciplines with the largest number of students and emphasis on clinical training-professional clinical psychology (PsyD) and social work-had the largest percentage of programs (67.3% and 61.7%, respectively) not requiring a didactic and clinical supervision in any EBT. There is a

  17. Film Program Notes from the Current Holdings of the Anthology Film Archives; Outlines of 41 Films.

    ERIC Educational Resources Information Center

    Anthology Film Archives, New York, NY.

    This collection of film program notes includes mixed commentary on some of the films held in the Anthology Film Archives (a film and book library in New York City). Some of the films are described by synopsis of the episodes and others by translation into English of the foreign language subtitles. However, each film noted is identified by full…

  18. Concordance of coronary artery calcium estimates between MDCT and electron beam tomography.

    PubMed

    Daniell, Anthony L; Wong, Nathan D; Friedman, John D; Ben-Yosef, Nachum; Miranda-Peats, Romalisa; Hayes, Sean W; Kang, Xingping; Sciammarella, Maria G; de Yang, Ling; Germano, Guido; Berman, Daniel S

    2005-12-01

    The objective of our study was to compare MDCT with electron beam tomography (EBT) for the quantification of coronary artery calcification (CAC). Sixty-eight patients underwent both MDCT and EBT within 2 months for the quantification of CAC. The images were scored in a blinded fashion and independently by two observers with a minimum of 7 days between the interpretations of images obtained from one scanner type to the other. Presence versus absence of CAC was discordant by EBT versus MDCT in 6% (n = 4) of the cases by observer 1, with one of these cases also discordant by observer 2. All cases except one (aortic calcium misidentified as CAC) were among those with a mean Agatston score of less than 5 present on EBT but absent on MDCT. EBT and MDCT scores correlated well (r = 0.98-0.99). The relative median variability between EBT and MDCT for the Agatston score was 24% for observer 1 and 27% for observer 2 and was 18% and 14%, respectively, for volume score (average for both observers: 27% for Agatston score and 16% for volume score). Scores were higher for EBT than MDCT in approximately half of the cases, with little systematic difference between the two (median EBT-MDCT difference: Agatston score, -0.55; volume score, 3.4 mm3). The absolute median difference averaged for the two observers was 28.75 for the Agatston score and 15.4 mm3 for the volume score. Differences in CAC measurements using EBT and MDCT are similar to interscan differences in CAC measurements previously reported for EBT or for other MDCT scanners individually.

  19. Lay Health Worker Involvement in Evidence-Based Treatment Delivery: A Conceptual Model to Address Disparities in Care.

    PubMed

    Barnett, Miya L; Lau, Anna S; Miranda, Jeanne

    2018-05-07

    Mobilizing lay health workers (LHWs) to deliver evidence-based treatments (EBTs) is a workforce strategy to address mental health disparities in underserved communities. LHWs can be leveraged to support access to EBTs in a variety of ways, from conducting outreach for EBTs delivered by professional providers to serving as the primary treatment providers. This critical review provides an overview of how LHW-supported or -delivered EBTs have been leveraged in low-, middle-, and high-income countries (HICs). We propose a conceptual model for LHWs to address drivers of service disparities, which relate to the overall supply of the EBTs provided and the demand for these treatments. The review provides illustrative case examples that demonstrate how LHWs have been leveraged globally and domestically to increase access to mental health services. It also discusses challenges and recommendations regarding implementing LHW-supported or -delivered EBTs.

  20. Total body irradiation with a compensator fabricated using a 3D optical scanner and a 3D printer

    NASA Astrophysics Data System (ADS)

    Park, So-Yeon; Kim, Jung-in; Joo, Yoon Ha; Lee, Jung Chan; Park, Jong Min

    2017-05-01

    We propose bilateral total body irradiation (TBI) utilizing a 3D printer and a 3D optical scanner. We acquired surface information of an anthropomorphic phantom with the 3D scanner and fabricated the 3D compensator with the 3D printer, which could continuously compensate for the lateral missing tissue of an entire body from the beam’s eye view. To test the system’s performance, we measured doses with optically stimulated luminescent dosimeters (OSLDs) as well as EBT3 films with the anthropomorphic phantom during TBI without a compensator, conventional bilateral TBI, and TBI with the 3D compensator (3D TBI). The 3D TBI showed the most uniform dose delivery to the phantom. From the OSLD measurements of the 3D TBI, the deviations between the measured doses and the prescription dose ranged from  -6.7% to 2.4% inside the phantom and from  -2.3% to 0.6% on the phantom’s surface. From the EBT3 film measurements, the prescription dose could be delivered to the entire body of the phantom within  ±10% accuracy, except for the chest region, where tissue heterogeneity is extreme. The 3D TBI doses were much more uniform than those of the other irradiation techniques, especially in the anterior-to-posterior direction. The 3D TBI was advantageous, owing to its uniform dose delivery as well as its efficient treatment procedure.

  1. Dose-current discharge correlation analysis in a Mather type Plasma Focus device for medical applications

    NASA Astrophysics Data System (ADS)

    Sumini, M.; Mostacci, D.; Tartari, A.; Mazza, A.; Cucchi, G.; Isolan, L.; Buontempo, F.; Zironi, I.; Castellani, G.

    2017-11-01

    In a Plasma Focus device the plasma collapses into the pinch where it reaches thermonuclear conditions for a few tens of nanoseconds, becoming a multi-radiation source. The nature of the radiation generated depends on the gas filling the chamber and the device working parameters. The self-collimated electron beam generated in the backward direction with respect to the plasma motion is one of the main radiation sources of interest also for medical applications. The electron beam may be guided against a high Z material target to produce an X-ray beam. This technique offers an ultra-high dose rate source of X-rays, able to deliver during the pinch a massive dose (up to 1 Gy per discharge for the PFMA-3 test device), as measured with EBT3 GafchromicⒸfilm tissue equivalent dosimeters. Given the stochastic behavior of the discharge process, a reliable on-line estimate of the dose-delivered is a very challenging task, in some way preventing a systematic application as a potentially interesting therapy device. This work presents an approach to linking the dose registered by the EBT3 GafchromicⒸfilms with the information contained in the signal recorded during the current discharge process. Processing the signal with the Wigner-Ville distribution, a spectrogram was obtained, displaying the information on intensity at various frequency scales, identifying the band of frequencies representative of the pinch events and define some patterns correlated with the dose.

  2. The National Film Registry: Acquiring Our Film Heritage.

    ERIC Educational Resources Information Center

    Ziegler, Roy A.

    The National Film Registry, which is primarily a designated list of films to be preserved by the Library of Congress, is also a valuable tool for selecting "films that are culturally, historically, and aesthetically significant." Following a brief discussion of the history and selection process of the National Film Registry, Southeast…

  3. Increasing supplemental nutrition assistance program/electronic benefits transfer sales at farmers' markets with vendor-operated wireless point-of-sale terminals.

    PubMed

    Buttenheim, Alison M; Havassy, Joshua; Fang, Michelle; Glyn, Jonathan; Karpyn, Allison E

    2012-05-01

    Supplemental Nutrition Assistance Program (SNAP) (formerly Food Stamp Program) participants can use their benefits at many farmers' markets. However, most markets have only one market-operated wireless point-of-sale (POS) card swipe terminal for electronic benefits transfer (EBT) transactions. It is not known whether providing each farmer/vendor with individual wireless POS terminals and subsidizing EBT fees will increase SNAP/EBT purchases at farmers' markets. To evaluate the effects of multiple vendor-operated wireless POS terminals (vs a single market-operated terminal) on use of SNAP benefits at an urban farmers' market. Time-series analyses of SNAP/EBT sales. The Clark Park farmers' market in West Philadelphia, PA, which accounts for one quarter of all SNAP/EBT sales at farmers' markets in Pennsylvania. Vendors were provided with individual wireless POS terminals for 9 months (June 2008-February 2009.) The pilot program covered all equipment and wireless service costs and transaction fees associated with SNAP/EBT, credit, and debit sales. Monthly SNAP/EBT sales at the Clark Park farmers' market. SNAP/EBT sales data were collected for 48 months (January 2007-December 2010). Time-series regression analysis was used to estimate the effect of the intervention period (June 2008-February 2009) on SNAP/EBT sales, controlling for seasonal effects and total SNAP benefits issued in Philadelphia. The intervention was associated with a 38% increase in monthly SNAP/EBT sales. Effects were greatest during the busy fall market seasons. SNAP/EBT sales did not remain significantly higher after the intervention period. Providing individual wireless POS terminals to farmers' market vendors leads to increased sales. However, market vendors indicated that subsidies for equipment costs and fees would be needed to break even. Currently, SNAP provides some support for these services for supermarket and other SNAP retailers with landline access, but not for farmers' markets. Copyright

  4. Clinical Decision-Making in Community Children's Mental Health: Using Innovative Methods to Compare Clinicians With and Without Training in Evidence-Based Treatment.

    PubMed

    Baker-Ericzén, Mary J; Jenkins, Melissa M; Park, Soojin; Garland, Ann F

    2015-02-01

    Mental health professionals' decision-making practice is an area of increasing interest and importance, especially in the pediatric research and clinical communities. The present study explored the role of prior training in evidence-based treatments on clinicians' assessment and treatment formulations using case vignettes. Specifically, study aims included using the Naturalistic Decision Making (NDM) cognitive theory to 1) examine potential associations between EBT training and decision-making processes (novice versus expert type), and 2) explore how client and family contextual information affects clinical decision-making. Forty-eight clinicians across two groups (EBT trained=14, Not EBT trained=34) participated. Clinicians were comparable on professional experience, demographics, and discipline. The quasi-experimental design used an analog "think aloud" method where clinicians read case vignettes about a child with disruptive behavior problems and verbalized case conceptualization and treatment planning out-loud. Responses were coded according to NDM theory. MANOVA results were significant for EBT training status such that EBT trained clinicians' displayed cognitive processes more closely aligned with "expert" decision-makers and non-EBT trained clinicians' decision processes were more similar to "novice" decision-makers, following NDM theory. Non-EBT trained clinicians assigned significantly more diagnoses, provided less detailed treatment plans and discussed fewer EBTs. Parent/family contextual information also appeared to influence decision-making. This study offers a preliminary investigation of the possible broader impacts of EBT training and potential associations with development of expert decision-making skills. Targeting clinicians' decision-making may be an important avenue to pursue within dissemination-implementation efforts in mental health practice.

  5. Film Reviews.

    ERIC Educational Resources Information Center

    Lance, Larry M.; Atwater, Lynn

    1987-01-01

    Reviews four Human Sexuality films and videos. These are: "Personal Decisions" (Planned Parenthood Federation of America, 1985); "The Touch Film" (Sterling Production, 1986); "Rethinking Rape" (Film Distribution Center, 1985); "Not A Love Story" (National Film Board of Canada, 1981). (AEM)

  6. Film and membrane-model thermodynamics of free thin liquid films.

    PubMed

    Radke, C J

    2015-07-01

    In spite of over 7 decades of effort, the thermodynamics of thin free liquid films (as in emulsions and foams) lacks clarity. Following a brief review of the meaning and measurement of thin-film forces (i.e., conjoining/disjoining pressures), we offer a consistent analysis of thin-film thermodynamics. By carefully defining film reversible work, two distinct thermodynamic formalisms emerge: a film model with two zero-volume membranes each of film tension γ(f) and a membrane model with a single zero-volume membrane of membrane tension 2γ(m). In both models, detailed thermodynamic analysis gives rise to thin-film Gibbs adsorption equations that allow calculation of film and membrane tensions from measurements of disjoining-pressure isotherms. A modified Young-Laplace equation arises in the film model to calculate film-thickness profiles from the film center to the surrounding bulk meniscus. No corresponding relation exists in the membrane model. Illustrative calculations of disjoining-pressure isotherms for water are presented using square-gradient theory. We report considerable deviations from Hamaker theory for films less than about 3 nm in thickness. Such thin films are considerably more attractive than in classical Hamaker theory. Available molecular simulations reinforce this finding. Copyright © 2014 Elsevier Inc. All rights reserved.

  7. SU-F-T-32: Evaluation of the Performance of a Multiple-Array-Diode Detector for Quality Assurance Tests in High-Dose-Rate Brachytherapy with Ir-192 Source

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Harpool, K; De La Fuente Herman, T; Ahmad, S

    Purpose: To evaluate the performance of a two-dimensional (2D) array-diode- detector for geometric and dosimetric quality assurance (QA) tests of high-dose-rate (HDR) brachytherapy with an Ir-192-source. Methods: A phantom setup was designed that encapsulated a two-dimensional (2D) array-diode-detector (MapCheck2) and a catheter for the HDR brachytherapy Ir-192 source. This setup was used to perform both geometric and dosimetric quality assurance for the HDR-Ir192 source. The geometric tests included: (a) measurement of the position of the source and (b) spacing between different dwell positions. The dosimteric tests include: (a) linearity of output with time, (b) end effect and (c) relative dosemore » verification. The 2D-dose distribution measured with MapCheck2 was used to perform the previous tests. The results of MapCheck2 were compared with the corresponding quality assurance testes performed with Gafchromic-film and well-ionization-chamber. Results: The position of the source and the spacing between different dwell-positions were reproducible within 1 mm accuracy by measuring the position of maximal dose using MapCheck2 in contrast to the film which showed a blurred image of the dwell positions due to limited film sensitivity to irradiation. The linearity of the dose with dwell times measured from MapCheck2 was superior to the linearity measured with ionization chamber due to higher signal-to-noise ratio of the diode readings. MapCheck2 provided more accurate measurement of the end effect with uncertainty < 1.5% in comparison with the ionization chamber uncertainty of 3%. Although MapCheck2 did not provide absolute calibration dosimeter for the activity of the source, it provided accurate tool for relative dose verification in HDR-brachytherapy. Conclusion: The 2D-array-diode-detector provides a practical, compact and accurate tool to perform quality assurance for HDR-brachytherapy with an Ir-192 source. The diodes in MapCheck2 have high radiation

  8. End-to-end test of spatial accuracy in Gamma Knife treatments for trigeminal neuralgia

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Brezovich, Ivan A., E-mail: ibrezovich@uabmc.edu; Wu, Xingen; Duan, Jun

    2014-11-01

    Purpose: Spatial accuracy is most crucial when small targets like the trigeminal nerve are treated. Although current quality assurance procedures typically verify that individual apparatus, like the MRI scanner, CT scanner, Gamma Knife, etc., are meeting specifications, the cumulative error of all equipment and procedures combined may exceed safe margins. This study uses an end-to-end approach to assess the overall targeting errors that may have occurred in individual patients previously treated for trigeminal neuralgia. Methods: The trigeminal nerve is simulated by a 3 mm long, 3.175 mm (1/8 in.) diameter MRI-contrast filled cavity embedded within a PMMA plastic capsule. Themore » capsule is positioned within the head frame such that the location of the cavity matches the Gamma Knife coordinates of an arbitrarily chosen, previously treated patient. Gafchromic EBT2 film is placed at the center of the cavity in coronal and sagittal orientations. The films are marked with a pinprick to identify the cavity center. Treatments are planned for radiation delivery with 4 mm collimators according to MRI and CT scans using the clinical localizer boxes and acquisition protocols. Shots are planned so that the 50% isodose surface encompasses the cavity. Following irradiation, the films are scanned and analyzed. Targeting errors are defined as the distance between the pinprick, which represents the intended target, and the centroid of the 50% isodose line, which is the center of the radiation field that was actually delivered. Results: Averaged over ten patient simulations, targeting errors along the x, y, and z coordinates (patient’s left-to-right, posterior-to-anterior, and head-to-foot) were, respectively, −0.060 ± 0.363, −0.350 ± 0.253, and 0.348 ± 0.204 mm when MRI was used for treatment planning. Planning according to CT exhibited generally smaller errors, namely, 0.109 ± 0.167, −0.191 ± 0.144, and 0.211 ± 0.094 mm. The largest errors along individual axes

  9. Surface dose measurements with commonly used detectors: a consistent thickness correction method

    PubMed Central

    Higgins, Patrick

    2015-01-01

    The purpose of this study was to review application of a consistent correction method for the solid state detectors, such as thermoluminescent dosimeters (chips (cTLD) and powder (pTLD)), optically stimulated detectors (both closed (OSL) and open (eOSL)), and radiochromic (EBT2) and radiographic (EDR2) films. In addition, to compare measured surface dose using an extrapolation ionization chamber (PTW 30‐360) with other parallel plate chambers RMI‐449 (Attix), Capintec PS‐033, PTW 30‐329 (Markus) and Memorial. Measurements of surface dose for 6 MV photons with parallel plate chambers were used to establish a baseline. cTLD, OSLs, EDR2, and EBT2 measurements were corrected using a method which involved irradiation of three dosimeter stacks, followed by linear extrapolation of individual dosimeter measurements to zero thickness. We determined the magnitude of correction for each detector and compared our results against an alternative correction method based on effective thickness. All uncorrected surface dose measurements exhibited overresponse, compared with the extrapolation chamber data, except for the Attix chamber. The closest match was obtained with the Attix chamber (−0.1%), followed by pTLD (0.5%), Capintec (4.5%), Memorial (7.3%), Markus (10%), cTLD (11.8%), eOSL (12.8%), EBT2 (14%), EDR2 (14.8%), and OSL (26%). Application of published ionization chamber corrections brought all the parallel plate results to within 1% of the extrapolation chamber. The extrapolation method corrected all solid‐state detector results to within 2% of baseline, except the OSLs. Extrapolation of dose using a simple three‐detector stack has been demonstrated to provide thickness corrections for cTLD, eOSLs, EBT2, and EDR2 which can then be used for surface dose measurements. Standard OSLs are not recommended for surface dose measurement. The effective thickness method suffers from the subjectivity inherent in the inclusion of measured percentage depth‐dose curves

  10. Surface dose measurements with commonly used detectors: a consistent thickness correction method.

    PubMed

    Reynolds, Tatsiana A; Higgins, Patrick

    2015-09-08

    The purpose of this study was to review application of a consistent correction method for the solid state detectors, such as thermoluminescent dosimeters (chips (cTLD) and powder (pTLD)), optically stimulated detectors (both closed (OSL) and open (eOSL)), and radiochromic (EBT2) and radiographic (EDR2) films. In addition, to compare measured surface dose using an extrapolation ionization chamber (PTW 30-360) with other parallel plate chambers RMI-449 (Attix), Capintec PS-033, PTW 30-329 (Markus) and Memorial. Measurements of surface dose for 6MV photons with parallel plate chambers were used to establish a baseline. cTLD, OSLs, EDR2, and EBT2 measurements were corrected using a method which involved irradiation of three dosimeter stacks, followed by linear extrapolation of individual dosimeter measurements to zero thickness. We determined the magnitude of correction for each detector and compared our results against an alternative correction method based on effective thickness. All uncorrected surface dose measurements exhibited overresponse, compared with the extrapolation chamber data, except for the Attix chamber. The closest match was obtained with the Attix chamber (-0.1%), followed by pTLD (0.5%), Capintec (4.5%), Memorial (7.3%), Markus (10%), cTLD (11.8%), eOSL (12.8%), EBT2 (14%), EDR2 (14.8%), and OSL (26%). Application of published ionization chamber corrections brought all the parallel plate results to within 1% of the extrapolation chamber. The extrapolation method corrected all solid-state detector results to within 2% of baseline, except the OSLs. Extrapolation of dose using a simple three-detector stack has been demonstrated to provide thickness corrections for cTLD, eOSLs, EBT2, and EDR2 which can then be used for surface dose measurements. Standard OSLs are not recommended for surface dose measurement. The effective thickness method suffers from the subjectivity inherent in the inclusion of measured percentage depth-dose curves and is not

  11. Measuring Practitioner Attitudes toward Evidence-Based Treatments: A Validation Study

    ERIC Educational Resources Information Center

    Ashcraft, Rindee G. P.; Foster, Sharon L.; Lowery, Amy E.; Henggeler, Scott W.; Chapman, Jason E.; Rowland, Melisa D.

    2011-01-01

    A better understanding of clinicians' attitudes toward evidence-based treatments (EBT) will presumably enhance the transfer of EBTs for substance-abusing adolescents from research to clinical application. The reliability and validity of two measures of therapist attitudes toward EBT were examined: the Evidence-Based Practice Attitude Scale…

  12. SU-F-P-47: Estimation of Skin Dose by Performing the Measurements On Cylindrical Phantom

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Bosma, S; Sanders, M; Aryal, P

    Purpose: To evaluate the skin dose by performing the measurements on cylindrical phantom with 6X beam. Methods: A cylindrical phantom was used to best model a patient surface. The source to surface distance (SSD) was 100 cm at phantom surface along central axis (CAX). The EBT2 films were cut into 2×2 cm2 pieces. Each piece of film was placed at CAX on phantom surface for each measurement at 0°, 15°, 30°, 45°, 60°, 75°, and 90° gantry angles for field sizes of 5×5, 10×10, 15×15, and 20×20 cm{sup 2} respectively. One hundred monitor units (MU) with 6X beam were deliveredmore » for each set up. Similarly, the measurements were repeated using lithium fluoride (LiF) thermoluminescent dosimeter (TLD) chips (1X1X1 mm{sup 3}). Two TLD chips were placed for each gantry angle and field size. The calibration curves were produced for both film and TLD. The computed tomography (CT) was also performed on the same cylindrical phantom and dose was evaluated at the phantom surface using Eclipse treatment planning system ( AAA algorithm) for skin dose comparison. Results: Data showed small differences at smaller angles among EBT2, TLD and Eclipse treatment planning system. But Eclipse treatment planning system under estimated the skin dose between 20% and 50% at larger gantry angles (between 40° and 80°) at all field sizes before dose differences began to converge. Conclusion: Given this data, we can conclude that Eclipse treatment planning system under estimated the dose especially between 40 and 80 degrees of obliquity compared to the measurements results. Ideally, this study can be applied largely to head and neck patients where contours differ drastically and where skin dose is paramount.« less

  13. Single-layer nano-carbon film, diamond film, and diamond/nano-carbon composite film field emission performance comparison

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wang, Xiaoping, E-mail: wxpchina64@aliyun.com, E-mail: wxpchina@sohu.com; Shanghai Key Laboratory of Modern Optical System, Shanghai 200093; Wang, Jinye

    A series of single-layer nano-carbon (SNC) films, diamond films, and diamond/nano-carbon (D/NC) composite films have been prepared on the highly doped silicon substrate by using microwave plasma chemical vapor deposition techniques. The films were characterised by scanning electron microscopy, Raman spectroscopy, and field emission I-V measurements. The experimental results indicated that the field emission maximum current density of D/NC composite films is 11.8–17.8 times that of diamond films. And the field emission current density of D/NC composite films is 2.9–5 times that of SNC films at an electric field of 3.0 V/μm. At the same time, the D/NC composite film exhibitsmore » the advantage of improved reproducibility and long term stability (both of the nano-carbon film within the D/NC composite cathode and the SNC cathode were prepared under the same experimental conditions). And for the D/NC composite sample, a high current density of 10 mA/cm{sup 2} at an electric field of 3.0 V/μm was obtained. Diamond layer can effectively improve the field emission characteristics of nano-carbon film. The reason may be due to the diamond film acts as the electron acceleration layer.« less

  14. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Cebe, M; Pacaci, P; Mabhouti, H

    Purpose: In this study, the two available calculation algorithms of the Varian Eclipse treatment planning system(TPS), the electron Monte Carlo(eMC) and General Gaussian Pencil Beam(GGPB) algorithms were used to compare measured and calculated peripheral dose distribution of electron beams. Methods: Peripheral dose measurements were carried out for 6, 9, 12, 15, 18 and 22 MeV electron beams of Varian Triology machine using parallel plate ionization chamber and EBT3 films in the slab phantom. Measurements were performed for 6×6, 10×10 and 25×25cm{sup 2} cone sizes at dmax of each energy up to 20cm beyond the field edges. Using the same filmmore » batch, the net OD to dose calibration curve was obtained for each energy. Films were scanned 48 hours after irradiation using an Epson 1000XL flatbed scanner. Dose distribution measured using parallel plate ionization chamber and EBT3 film and calculated by eMC and GGPB algorithms were compared. The measured and calculated data were then compared to find which algorithm calculates peripheral dose distribution more accurately. Results: The agreement between measurement and eMC was better than GGPB. The TPS underestimated the out of field doses. The difference between measured and calculated doses increase with the cone size. The largest deviation between calculated and parallel plate ionization chamber measured dose is less than 4.93% for eMC, but it can increase up to 7.51% for GGPB. For film measurement, the minimum gamma analysis passing rates between measured and calculated dose distributions were 98.2% and 92.7% for eMC and GGPB respectively for all field sizes and energies. Conclusion: Our results show that the Monte Carlo algorithm for electron planning in Eclipse is more accurate than previous algorithms for peripheral dose distributions. It must be emphasized that the use of GGPB for planning large field treatments with 6 MeV could lead to inaccuracies of clinical significance.« less

  15. Clinical Decision-Making in Community Children’s Mental Health: Using Innovative Methods to Compare Clinicians With and Without Training in Evidence-Based Treatment

    PubMed Central

    Baker-Ericzén, Mary J.; Jenkins, Melissa M.; Park, Soojin; Garland, Ann F.

    2014-01-01

    Background Mental health professionals’ decision-making practice is an area of increasing interest and importance, especially in the pediatric research and clinical communities. Objective The present study explored the role of prior training in evidence-based treatments on clinicians’ assessment and treatment formulations using case vignettes. Specifically, study aims included using the Naturalistic Decision Making (NDM) cognitive theory to 1) examine potential associations between EBT training and decision-making processes (novice versus expert type), and 2) explore how client and family contextual information affects clinical decision-making. Methods Forty-eight clinicians across two groups (EBT trained=14, Not EBT trained=34) participated. Clinicians were comparable on professional experience, demographics, and discipline. The quasi-experimental design used an analog “think aloud” method where clinicians read case vignettes about a child with disruptive behavior problems and verbalized case conceptualization and treatment planning out-loud. Responses were coded according to NDM theory. Results MANOVA results were significant for EBT training status such that EBT trained clinicians’ displayed cognitive processes more closely aligned with “expert” decision-makers and non-EBT trained clinicians’ decision processes were more similar to “novice” decision-makers, following NDM theory. Non-EBT trained clinicians assigned significantly more diagnoses, provided less detailed treatment plans and discussed fewer EBTs. Parent/family contextual information also appeared to influence decision-making. Conclusion This study offers a preliminary investigation of the possible broader impacts of EBT training and potential associations with development of expert decision-making skills. Targeting clinicians’ decision-making may be an important avenue to pursue within dissemination-implementation efforts in mental health practice. PMID:25892901

  16. Single‐fraction spine SBRT end‐to‐end testing on TomoTherapy, Vero, TrueBeam, and CyberKnife treatment platforms using a novel anthropomorphic phantom

    PubMed Central

    Kaufman, Isaac; Powell, Rachel; Pandya, Shalini; Somnay, Archana; Bossenberger, Todd; Ramirez, Ezequiel; Reynolds, Robert; Solberg, Timothy; Burmeister, Jay

    2015-01-01

    Spine SBRT involves the delivery of very high doses of radiation to targets adjacent to the spinal cord and is most commonly delivered in a single fraction. Highly conformal planning and accurate delivery of such plans is imperative for successful treatment without catastrophic adverse effects. End–to‐end testing is an important practice for evaluating the entire treatment process from simulation through treatment delivery. We performed end‐to‐end testing for a set of representative spine targets planned and delivered using four different treatment planning systems (TPSs) and delivery systems to evaluate the various capabilities of each. An anthropomorphic E2E SBRT phantom was simulated and treated on each system to evaluate agreement between measured and calculated doses. The phantom accepts ion chambers in the thoracic region and radiochromic film in the lumbar region. Four representative targets were developed within each region (thoracic and lumbar) to represent different presentations of spinal metastases and planned according to RTOG 0631 constraints. Plans were created using the TomoTherapy TPS for delivery using the Hi·Art system, the iPlan TPS for delivery using the Vero system, the Eclipse TPS for delivery using the TrueBeam system in both flattened and flattening filter free (FFF), and the MultiPlan TPS for delivery using the CyberKnife system. Delivered doses were measured using a 0.007 cm3 ion chamber in the thoracic region and EBT3 GAFCHROMIC film in the lumbar region. Films were scanned and analyzed using an Epson Expression 10000XL flatbed scanner in conjunction with FilmQAPro2013. All treatment platforms met all dose constraints required by RTOG 0631. Ion chamber measurements in the thoracic targets delivered an overall average difference of 1.5%. Specifically, measurements agreed with the TPS to within 2.2%, 3.2%, 1.4%, 3.1%, and 3.0% for all three measureable cases on TomoTherapy, Vero, TrueBeam (FFF), TrueBeam (flattened), and Cyber

  17. A comparison of a behavioral weight loss program to a stress management program: A pilot randomized controlled trial.

    PubMed

    Webber, Kelly H; Casey, Erin M; Mayes, Lindsey; Katsumata, Yuriko; Mellin, Laurel

    2016-01-01

    This study compared a behavioral weight loss program (BWL) with a stress management-based program, Emotional Brain Training (EBT), on weight loss, blood pressure, depression, perceived stress, diet, and physical activity. Subjects with a body mass index (BMI) of >28 and <45 kg/m(2) were recruited in Lexington, Kentucky in January 2014 and randomized to BWL or EBT for a 20-week intervention. Of those recruited, 49 participants were randomized to EBT or BWL. Randomization and allocation to group were performed using SPSS software. Weight, blood pressure, depression, perceived stress, dietary intake, and physical activity were measured at baseline, 10 week, and 20 week. Linear models for change over time were fit to calculate 95% confidence intervals of intervention effects. BWL produced greater changes in BMI than EBT at both 10 (P = 0.02) and 20 wk (P = 0.03). At 10 wk, both EBT and BWL improved BMI, systolic blood pressure, depression and perceived stress (P < 0.05). BWL also improved diastolic blood pressure (P = 0.005). At 20 wk, EBT maintained improvements in BMI, systolic blood pressure, depression, and perceived stress while BWL maintained improvements only in BMI and depression (P < 0.05). BWL produced greater weight loss than EBT; however, EBT produced sustained improvements in stress, depression, and systolic blood pressure. A combination of the two approaches should be explored. Copyright © 2016 Elsevier Inc. All rights reserved.

  18. Film and History.

    ERIC Educational Resources Information Center

    Schaber, Robin L.

    2002-01-01

    Provides an annotated bibliography of Web sites that focus on using film to teach history. Includes Web sites in five areas: (1) film and education; (2) history of cinema; (3) film and history resources; (4) film and women; and (5) film organizations. (CMK)

  19. Drop dynamics on a thin film: Thin film rupture

    NASA Astrophysics Data System (ADS)

    Carlson, Andreas; Kim, Pilnam; Stone, Howard A.

    2011-11-01

    The spreading of a water drop on an oil film that covers a solid substrate is a common event in many industrial processes. We study in experiments the dynamics of a water drop on a thin silicone oil film and quantify its interaction with the solid substrate that supports the film. The oil film becomes unstable and ruptures for solids that are hydrophilic. We determine the ``waiting time,'' the time it takes the water drop to drain the silicone film. This timescale is found to highly depend on how well water wets the solid, illustrating the interplay between intermolecular and hydrodynamic forces in the phenomenon. A phase diagram for the thin film stability is extracted based on waters equilibrium contact angle on the solid, which shows that we can either promote or inhibit de-wetting. As water comes in direct contact with the solid, it spreads and peels off the silicone film. We show the influence of viscosity, equilibrium contact angle and film height on the opening radius of the hole formed as the solid de-wets.

  20. Contemporary Films' Mini Course on Film Study.

    ERIC Educational Resources Information Center

    Schillaci, Peter

    This minicourse on film study can be a unit in English, in arts, or in the humanities. It can help to launch a film study course or complement an introduction to theater. Whatever form it takes, it helps to build a bridge to the student's media environment. Part one, the language of images, utilizes four films which demonstrate the basic elements…

  1. Dietary Behaviors among Public Health Center Clients with Electronic Benefit Transfer Access at Farmers' Markets.

    PubMed

    Robles, Brenda; Montes, Christine E; Nobari, Tabashir Z; Wang, May C; Kuo, Tony

    2017-01-01

    Although increasing access to electronic benefit transfer (EBT) at farmers' markets has become a popular strategy for encouraging healthy eating, its relationships to a number of dietary behaviors in low-income populations are not well understood. To describe the frequency of and relationships between EBT access, fruit and vegetable intake, and sugar-sweetened beverage (SSB) consumption among public health center (PHC) clients with access to EBT at farmers' markets during 2011-2012. Cross-sectional. Low-income participants recruited from the waiting rooms of five multipurpose PHCs operated by the Los Angeles County Department of Public Health. Fruit and vegetable and SSB consumption (number per week). Data from the 2012 Los Angeles County Health and Nutrition Examination Survey were analyzed using multivariable regressions, with EBT access at farmers' markets as the primary independent variable. Covariates included EBT use, transportation behaviors, neighborhood attributes, and sociodemographic characteristics. A total of 1,503 adults participated in the survey (response rate=69%). Of these, 529 reported receiving EBT benefits. Among these benefits recipients, 64% were women, 54% were aged 25 to 44 years, 62% were black, and 75% were unemployed or part-time employed. In multivariable regression analyses, EBT access at farmers' markets was positively associated with higher fruit and vegetable consumption; however, an association to SSB consumption was not demonstrated. EBT access at farmers' markets is related to higher fruit and vegetable consumption among PHC clients in Los Angeles County. However, the finding of no association to SSB consumption raises important questions about the need for strategies to discourage EBT recipients' purchase of foods of minimal nutritional value in other venues that accept nutrition assistance program benefits. Copyright © 2017 Academy of Nutrition and Dietetics. Published by Elsevier Inc. All rights reserved.

  2. Virtual reality exposure-based therapy for the treatment of post-traumatic stress disorder: a review of its efficacy, the adequacy of the treatment protocol, and its acceptability

    PubMed Central

    Botella, Cristina; Serrano, Berenice; Baños, Rosa M; Garcia-Palacios, Azucena

    2015-01-01

    Introduction The essential feature of post-traumatic stress disorder (PTSD) is the development of characteristic symptoms following exposure to one or more traumatic events. According to evidence-based intervention guidelines and empirical evidence, one of the most extensively researched and validated treatments for PTSD is prolonged exposure to traumatic events; however, exposure therapy can present some limitations. Virtual reality (VR) can help to improve prolonged exposure because it creates fictitious, safe, and controllable situations that can enhance emotional engagement and acceptance. Objective In addition to carrying out a review to evaluate the efficacy of VR exposure-based therapy (VR-EBT) for the treatment of PTSD, the aim of this study was to contribute to analyzing the use of VR-EBT by: first, evaluating the adequacy of psychological treatment protocols that use VR-EBT to treat PTSD; and second, analyzing the acceptability of VR-EBT. Method We performed a replica search with descriptors and databases used in two previous reviews and updated to April 2015. Next, we carried out an evaluation of the efficacy, adequacy, and acceptability of VR-EBT protocols. Results Results showed that VR-EBT was effective in the treatment of PTSD. The findings related to adequacy showed that not all studies using VR-EBT reported having followed the clinical guidelines for evidence-based interventions in the treatment of PTSD. Regarding acceptability, few studies evaluated this subject. However, the findings are very promising, and patients reported high acceptability and satisfaction with the inclusion of VR in the treatment of PTSD. Conclusion The main weaknesses identified in this review focus on the need for more controlled studies, the need to standardize treatment protocols using VR-EBT, and the need to include assessments of acceptability and related variables. Finally, this paper highlights some directions and future perspectives for using VR-EBT in PTSD

  3. American Film Genres: Approaches to a Critical Theory of Popular Film.

    ERIC Educational Resources Information Center

    Kaminsky, Stuart M.

    This book is divided into twelve sections and contains photographs from many of the films discussed. The introduction defines film genre and describes the general theories behind this book; "The Individual Film" analyzes the film "Little Caesar" as it relates to the genre of gangster films; "Comparative Forms"…

  4. Surface applicator calibration and commissioning of an electronic brachytherapy system for nonmelanoma skin cancer treatment.

    PubMed

    Rong, Yi; Welsh, James S

    2010-10-01

    The Xoft Axxent x-ray source has been used for treating nonmelanoma skin cancer since the surface applicators became clinically available in 2009. The authors report comprehensive calibration procedures for the electronic brachytherapy (eBx) system with the surface applicators. The Xoft miniature tube (model S700) generates 50 kVp low-energy x rays. The new surface applicators are available in four sizes of 10, 20, 35, and 50 mm in diameter. The authors' tests include measurements of dose rate, air-gap factor, output stability, depth dose verification, beam flatness and symmetry, and treatment planning with patient specific cutout factors. The TG-61 in-air method was used as a guideline for acquiring nominal dose-rate output at the skin surface. A soft x-ray parallel-plate chamber (PTW T34013) and electrometer was used for the output commissioning. GafChromic EBT films were used for testing the properties of the treatment fields with the skin applicators. Solid water slabs were used to verify the depth dose and cutout factors. Patients with basal cell or squamous cell carcinoma were treated with eBx using a calibrated Xoft system with the low-energy x-ray source and the skin applicators. The average nominal dose-rate output at the skin surface for the 35 mm applicator is 1.35 Gy/min with +/- 5% variation for 16 sources. The dose-rate output and stability (within +/- 5% variation) were also measured for the remaining three applicators. For the same source, the output variation is within 2%. The effective source-surface distance was calculated based on the air-gap measurements for four applicator sizes. The field flatness and symmetry are well within 5%. Percentage depth dose in water was provided by factory measurements and can be verified using solid water slabs. Treatment duration was calculated based on the nominal dose rate, the prescription fraction size, the depth dose percentage, and the cutout factor. The output factor needs to be measured for each case with

  5. SU-E-T-470: Beam Performance of the Radiance 330 Proton Therapy System

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Nazaryan, H; Nazaryan, V; Wang, F

    2014-06-01

    Purpose: The ProTom Radiance 330 proton radiotherapy system is a fully functional, compact proton radiotherapy system that provides advanced proton delivery capabilities. It supports three-dimensional beam scanning with energy and intensity modulation. A series of measurements have been conducted to characterize the beam performance of the first installation of the system at the McLaren Proton Therapy Center in Flint, Michigan. These measurements were part of the technical commissioning of the system. Select measurements and results are presented. Methods: The Radiance 330 proton beam energy range is 70–250 MeV for treatment, and up to 330 MeV for proton tomography and radiography.more » Its 3-D scanning capability, together with a small beam emittance and momentum spread, provides a highly efficient beam delivery. During the technical commissioning, treatment plans were created to deliver uniform maps at various energies to perform Gamma Index analysis. EBT3 Gafchromic films were irradiated using the Planned irradiation maps. Bragg Peak chamber was used to test the dynamic range during a scan in one layer for high (250 MeV) and Low (70 MeV) energies. The maximum and minimum range, range adjustment and modulation, distal dose falloff (80%–20%), pencil beam spot size, spot placement accuracy were also measured. The accuracy testing included acquiring images, image registration, receiving correction vectors and applying the corrections to the robotic patient positioner. Results: Gamma Index analysis of the Treatment Planning System (TPS) data vs. Measured data showed more than 90% of points within (3%, 3mm) for the maps created by the TPS. At Isocenter Beam Size (One sigma) < 3mm at highest energy (250 MeV) in air. Beam delivery was within 0.6 mm of the intended target at the entrance and the exit of the beam, through the phantom. Conclusion: The Radiance 330 Beam Performance Measurements have confirmed that the system operates as designed with excellent

  6. SU-E-T-672: Real-Time In Vivo Dosimeters Using LiPCDA and Optical Fibers

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rink, A; Jaffray, DA; Croteau, A

    2015-06-15

    Purpose: To investigate dosimeter prototypes made with lithium pentacosa-10,12-diynoate (LiPCDA, the material used in GafChromic EBT films) and optical fibers for their suitability in real-time in vivo measurements. Methods: The prototypes, made with 500 µm plastic optical fibers and 8.5 µm thickness of LiPCDA at fiber tip, were irradiated with a 6 MV beam. To investigate the efficacy of pre-irradiation calibration, the probes were irradiated to 2 Gy twice, with 5 minutes in between. Net optical density values (netOD) around the main absorbance peak were compared, and effect of correcting second measurement by the first was assessed. Ageing was assessedmore » by irradiating two prototypes to 2 Gy and comparing the netOD to that obtained for 15 prototypes from the same batch 12–14 months earlier. To measure angular dependence, the probes were pre-irradiated with beam perpendicular to fiber axis and then, 5 minutes later either perpendicular or parallel to fiber axes. The thickness-corrected netOD measurements were compared. Results: Standard deviation of netOD for probes of the same batch was measured to be 5–6%. When netOD was corrected for material thickness by using results from the first irradiation, the standard deviation decreased to 1.3%. This was comparable to the uncertainty in measurements observed with a single probe and is attributed to variations in light output, spectrometer noise and splitter-to-probe connection variations. Comparison of netOD values obtained a year apart failed to illustrate statistically significant decrease in sensitivity due to ageing (0.38 ± 0.03 and 0.3656 ± 0.0003). NetOD measured with MV beam parallel to fiber was within error of netOD measured with MV beam perpendicular to fiber. Conclusion: Current prototype construction allows for shelf life of at least one year. With material thickness corrected for, the prototypes can measure dose with an uncertainty below 2% at a given energy and dose rate. This work has been

  7. Experimental measurement of radiological penumbra associated with intermediate energy x-rays (1 MV) and small radiosurgery field sizes

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Keller, Brian M.; Beachey, David J.; Pignol, Jean-Philippe

    2007-10-15

    Stereotactic radiosurgery is used to treat intracranial lesions with a high degree of accuracy. At the present time, x-ray energies at or above Co-60 gamma rays are used. Previous Monte Carlo simulations have demonstrated that intermediate energy x-ray photons or IEPs (defined to be photons in the energy range of 0.2-1.2 MeV), combined with small field sizes, produce a reduced radiological penumbra leading to a sharper dose gradient, improved dose homogeneity and sparing of critical anatomy adjacent to the target volume. This hypothesis is based on the fact that, for small x-ray fields, a dose outside the treatment volume ismore » dictated mainly by the range of electrons set into motion by x-ray photons. The purpose of this work is: (1) to produce intermediate energy x rays using a detuned medical linear accelerator (2) to characterize the energy of this beam (3) to measure the radiological penumbra for IEPs and small fields to compare with that produced by 6 MV x rays or Co-60, and (4) to compare these experimental measurements with Monte Carlo computer simulations. The maximum photon energy of our IEP x-ray spectrum was measured to be 1.2 MeV. Gafchromic EBT films (ISP Technologies, Wayne, NJ) were irradiated and read using a novel digital microscopy imaging system with high spatial resolution. Under identical irradiation conditions the measured radiological penumbra widths (80%-20% distance), for field sizes ranging from 0.3x0.3 to 4.0x4.0 cm{sup 2}, varied from 0.3-0.77 mm (1.2 MV) and from 1.1-2.1 mm (6 MV). Even more dramatic were the differences found when comparing the 90%-10% or the 95%-5% widths, which are in fact more significant in radiotherapy. Monte Carlo simulations agreed well with the experimental findings. The reduction in radiological penumbra could be substantial for specific clinical situations such as in the treatment of an ocular melanoma abutting the macula or for the treatment of functional disorders such as trigeminal neuralgia (a

  8. Potential benefits of dosimetric VMAT tracking verified with 3D film measurements.

    PubMed

    Crijns, Wouter; Defraene, Gilles; Van Herck, Hans; Depuydt, Tom; Haustermans, Karin; Maes, Frederik; Van den Heuvel, Frank

    2016-05-01

    To evaluate three different plan adaptation strategies using 3D film-stack dose measurements of both focal boost and hypofractionated prostate VMAT treatments. The adaptation strategies (a couch shift, geometric tracking, and dosimetric tracking) were applied for three realistic intrafraction prostate motions. A focal boost (35 × 2.2 and 35 × 2.7 Gy) and a hypofractionated (5 × 7.25 Gy) prostate VMAT plan were created for a heterogeneous phantom that allows for internal prostate motion. For these plans geometric tracking and dosimetric tracking were evaluated by ionization chamber (IC) point dose measurements (zero-D) and measurements using a stack of EBT3 films (3D). The geometric tracking applied translations, rotations, and scaling of the MLC aperture in response to realistic prostate motions. The dosimetric tracking additionally corrected the monitor units to resolve variations due to difference in depth, tissue heterogeneity, and MLC-aperture. The tracking was based on the positions of four fiducial points only. The film measurements were compared to the gold standard (i.e., IC measurements) and the planned dose distribution. Additionally, the 3D measurements were converted to dose volume histograms, tumor control probability, and normal tissue complication probability parameters (DVH/TCP/NTCP) as a direct estimate of clinical relevance of the proposed tracking. Compared to the planned dose distribution, measurements without prostate motion and tracking showed already a reduced homogeneity of the dose distribution. Adding prostate motion further blurs the DVHs for all treatment approaches. The clinical practice (no tracking) delivered the dose distribution inside the PTV but off target (CTV), resulting in boost dose errors up to 10%. The geometric and dosimetric tracking corrected the dose distribution's position. Moreover, the dosimetric tracking could achieve the planned boost DVH, but not the DVH of the more homogeneously irradiated prostate. A drawback

  9. Film: An Introduction.

    ERIC Educational Resources Information Center

    Fell, John L.

    "Understanding Film," the opening section of this book, discusses perceptions of and responses to film and the way in which experiences with and knowledge of other media affect film viewing. The second section, "Film Elements," analyzes the basic elements of film: the use of space and time, the impact of editing, sound and color, and the effects…

  10. Film Makers On Film Making.

    ERIC Educational Resources Information Center

    Geduld, Harry M., Ed.

    This collection includes essays by and interviews with more than 30 film-makers, both classic and contemporary, on the subjects of their major interests and procedures in making films. The directors are: Louis Lumiere, Cecil Hepworth, Edwin S. Porter, Mack Sennett, David W. Griffith, Robert Flaherty, Charles Chaplin, Eric von Stroheim, Dziga…

  11. Democratizing Implementation and Innovation in Mental Health Care.

    PubMed

    Saxe, Glenn; Acri, Mary

    2017-03-01

    Improvements in the quality of mental health care in the United States depend on the successful implementation of evidence-based treatments (EBT's) in typical settings of care. Unfortunately, there is little evidence that EBT's are used in ways that would approximate their established fidelity standards in such settings. This article describes an approach to more successful implementation of EBT's via a collaborative process between intervention developers and intervention users (e.g. providers, administrators, consumers) called Lead-user Innovation. Lead-user Innovation democratizes the implementation process by integrating the expertise of lead-users in the delivery, adaptation, innovation and evaluation of EBT's.

  12. Methods for producing complex films, and films produced thereby

    DOEpatents

    Duty, Chad E.; Bennett, Charlee J. C.; Moon, Ji -Won; Phelps, Tommy J.; Blue, Craig A.; Dai, Quanqin; Hu, Michael Z.; Ivanov, Ilia N.; Jellison, Jr., Gerald E.; Love, Lonnie J.; Ott, Ronald D.; Parish, Chad M.; Walker, Steven

    2015-11-24

    A method for producing a film, the method comprising melting a layer of precursor particles on a substrate until at least a portion of the melted particles are planarized and merged to produce the film. The invention is also directed to a method for producing a photovoltaic film, the method comprising depositing particles having a photovoltaic or other property onto a substrate, and affixing the particles to the substrate, wherein the particles may or may not be subsequently melted. Also described herein are films produced by these methods, methods for producing a patterned film on a substrate, and methods for producing a multilayer structure.

  13. Thin Films

    NASA Astrophysics Data System (ADS)

    Khorshidi, Zahra; Bahari, Ali; Gholipur, Reza

    2014-11-01

    Effect of annealing temperature on the characteristics of sol-gel-driven Ta ax La(1- a) x O y thin film spin-coated on Si substrate as a high- k gate dielectric was studied. Ta ax La(1- a) x O y thin films with different amounts of a were prepared (as-prepared samples). X-ray diffraction measurements of the as-prepared samples indicated that Ta0.3 x La0.7 x Oy film had an amorphous structure. Therefore, Ta0.3 x La0.7 x O y film was chosen to continue the present studies. The morphology of Ta0.3 x La0.7 x O y films was studied using scanning electron microscopy and atomic force microscopy techniques. The obtained results showed that the size of grain boundaries on Ta0.3 x La0.7 x O y film surfaces was increased with increasing annealing temperature. Electrical and optical characterizations of the as-prepared and annealed films were investigated as a function of annealing temperature using capacitance-voltage ( C- V) and current density-voltage ( J- V) measurements and the Tauc method. The obtained results demonstrated that Ta0.3 x La0.7 x O y films had high dielectric constant (≈27), wide band gap (≈4.5 eV), and low leakage current density (≈10-6 A/cm2 at 1 V).

  14. The National Film Board's Studio D: Feminists Making Films for Peace.

    ERIC Educational Resources Information Center

    Taylor, Anita

    Studio D of the National Film Board of Canada, a women's film making unit established to make films by, for, and about women, has created a group of five films that effectively develop the argument that women can and must join the effort to bring peace to a nuclear world. The first of these peace films, "If You Love This Planet,"…

  15. Films Kids Like. A Catalog of Short Films For Children.

    ERIC Educational Resources Information Center

    Rice, Susan, Ed.

    Information on the Children's Film Theater and a list of short films are contained in this catalog. The first section reviews the purpose of the Children's Film Theater, its history and what has been learned as a result of its efforts. Details on the mechanics of showing films for children are also covered, including matters such as screening…

  16. Filming eugenics: teaching the history of eugenics through film.

    PubMed

    Ooten, Melissa; Trembanis, Sarah

    2007-01-01

    In teaching eugenics to undergraduate students and general public audiences, film should he considered as a provocative and fruitful medium that can generate important discussions about the intersections among eugenics, gender, class, race, and sexuality. This paper considers the use of two films, A Bill of Divorcement and The Lynchburg Story, as pedagogical tools for the history of eugenics. The authors provide background information on the films and suggestions for using the films to foster an active engagement with the historical eugenics movement.

  17. Use of evidence-based treatments in substance abuse treatment programs serving American Indian and Alaska Native communities.

    PubMed

    Novins, Douglas K; Croy, Calvin D; Moore, Laurie A; Rieckmann, Traci

    2016-04-01

    Research and health surveillance activities continue to document the substantial disparities in the impacts of substance abuse on the health of American Indian and Alaska Native (AI/AN) people. While Evidence-Based Treatments (EBTs) hold substantial promise for improving treatment for AI/ANs with substance use problems (as they do for non-AI/ANs), anecdotal reports suggest that their use is limited. In this study, we examine the awareness of, attitudes toward, and use of EBTs in substance abuse treatment programs serving AI/AN communities. Data are drawn from the first national survey of tribal substance abuse treatment programs. Clinicians or clinical administrators from 192 programs completed the survey. Participants were queried about their awareness of, attitudes toward, and use of 9 psychosocial and 3 medication EBTs. Cognitive Behavioral Therapy (82.2%), Motivational Interviewing (68.6%), and Relapse Prevention Therapy (66.8%) were the most commonly implemented psychosocial EBTs; medications for psychiatric comorbidity was the most commonly implemented medication treatment (43.2%). Greater EBT knowledge and use were associated with both program (e.g., funding) and staff (e.g., educational attainment) characteristics. Only two of the commonly implemented psychosocial EBTs (Motivational Interviewing and Relapse Prevention Therapy) were endorsed as culturally appropriate by a majority of programs that had implemented them (55.9% and 58.1%, respectively). EBT knowledge and use is higher in substance abuse treatment programs serving AI/AN communities than has been previously estimated. However, many users of these EBTs continue to have concerns about their cultural appropriateness, which likely limits their further dissemination. Copyright © 2016 Elsevier Ireland Ltd. All rights reserved.

  18. Nanocomposite films

    DOEpatents

    Mitlin, David; , Ophus, Colin; Evoy, Stephane; Radmilovic, Velimir; Mohammadi, Reza; Westra, Ken; Nelson-Fitzpatrick, Nathaniel; Lee, Zonghoon

    2010-07-20

    A thin-film composition of nanocrystal molybdenum in an amorphous metallic matrix may be formed by co-sputtering Mo with aluminum or nickel. NEMS cantilevers may be formed from the film. The films exhibit high nanoindentation hardness and a reduction in roughness and intrinsic stress, while maintaining resistivity in the metallic range.

  19. On Film

    ERIC Educational Resources Information Center

    Watts, Marty

    2006-01-01

    In this article, the author discusses the role of window films in enhancing indoor air quality in schools. Historically, window film has been used to reduce temperatures in buildings prone to overheating. Too much solar energy entering through windows makes occupants uncomfortable and air conditioning more costly. Film has been a simple solution…

  20. UV actinometer film

    NASA Technical Reports Server (NTRS)

    Coulbert, C. D.; Gupta, A.; Pitts, J.

    1980-01-01

    Cumulative UV radiation can be measured by low-cost polymer film that is unaffacted by visible light. Useful for virtually any surface, film can help paint and plastics manufacturers determine how well their products stand up against UV radiation. Actinometer film uses photochemically sensitive compound that changes its chemical composition in response to solar radiation. Extent of chemical conversion depends on length exposure and can be measured by examining film sample with spectrophotometer. Film can be exposed from several seconds up to month.

  1. Water-evaporation reduction by duplex films: application to the human tear film.

    PubMed

    Cerretani, Colin F; Ho, Nghia H; Radke, C J

    2013-09-01

    Water-evaporation reduction by duplex-oil films is especially important to understand the physiology of the human tear film. Secreted lipids, called meibum, form a duplex film that coats the aqueous tear film and purportedly reduces tear evaporation. Lipid-layer deficiency is correlated with the occurrence of dry-eye disease; however, in-vitro experiments fail to show water-evaporation reduction by tear-lipid duplex films. We review the available literature on water-evaporation reduction by duplex-oil films and outline the theoretical underpinnings of spreading and evaporation kinetics that govern behavior of these systems. A dissolution-diffusion model unifies the data reported in the literature and identifies dewetting of duplex films into lenses as a key challenge to obtaining significant evaporation reduction. We develop an improved apparatus for measuring evaporation reduction by duplex-oil films including simultaneous assessment of film coverage, stability, and temperature, all under controlled external mass transfer. New data reported in this study fit into the larger body of work conducted on water-evaporation reduction by duplex-oil films. Duplex-oil films of oxidized mineral oil/mucin (MOx/BSM), human meibum (HM), and bovine meibum (BM) reduce water evaporation by a dissolution-diffusion mechanism, as confirmed by agreement between measurement and theory. The water permeability of oxidized-mineral-oil duplex films agrees with those reported in the literature, after correction for the presence of mucin. We find that duplex-oil films of bovine and human meibum at physiologic temperature reduce water evaporation only 6-8% for a 100-nm film thickness pertinent to the human tear film. Comparison to in-vivo human tear-evaporation measurements is inconclusive because evaporation from a clean-water surface is not measured and because the mass-transfer resistance is not characterized. Copyright © 2013 Elsevier B.V. All rights reserved.

  2. Thin teflon-like films for MEMS: Film properties and reliability studies

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Smith, B.K.; LaVigne, G.; Sniegowski, J.J.

    1998-07-01

    This work presents film properties and initial reliability studies for thin Teflon-like films applied to a unique test vehicle, the Sandia-designed and fabricated microengine. Results on microengines coated with the film show a factor of three improvement in their lifetime and an order of magnitude reduction in the coefficient of friction when compared to uncoated samples. Coefficients Of Friction (COF) of 0.07 for the Teflon-like film and 1.0 for uncoated samples are extracted from models which match the measured behavior of working microengines. These films, deposited form a plasma source, exhibit the ability to penetrate into very narrow, deep channelsmore » common to many MEMS devices. For as-deposited film, both the refractive index at 1.4 and the contact angle with water at 108{degree} show the film to be very similar to bulk Teflon PTFE. Film stability as a function of temperature has been examined using Fourier Transformation Infrared (FTIR) spectroscopy. The film structure as observed by the fluorine-carbon (F-C) peak is stable up to 200 C, but starts decomposing above 250 C. Film composition has been examined using X-ray photoelectron spectroscopy (XPS) and is quite different for directly exposed surfaces compared with deep, narrow channels where the deposition process is diffusion limited.« less

  3. Film in Development.

    ERIC Educational Resources Information Center

    Information Center on Instructional Technology Report, 1976

    1976-01-01

    The positive and negative aspects of using film as a medium for inducing change in developing nations are explored in this issue. Critics of the use of film for development cite (1) the culture specific nature of film content of foreign produced films; (2) the technical, economic, and logistical problems of in-country production; and (3) lack of…

  4. Ag films deposited on Si and Ti: How the film-substrate interaction influences the nanoscale film morphology

    NASA Astrophysics Data System (ADS)

    Ruffino, F.; Torrisi, V.

    2017-11-01

    Submicron-thick Ag films were sputter deposited, at room temperature, on Si, covered by the native SiO2 layer, and on Ti, covered by the native TiO2 layer, under normal and oblique deposition angle. The aim of this work was to study the morphological differences in the grown Ag films on the two substrates when fixed all the other deposition parameters. In fact, the surface diffusivity of the Ag adatoms is different on the two substrates (higher on the SiO2 surface) due to the different Ag-SiO2 and Ag-TiO2 atomic interactions. So, the effect of the adatoms surface diffusivity, as determined by the adatoms-substrate interaction, on the final film morphology was analyzed. To this end, microscopic analyses were used to study the morphology of the grown Ag films. Even if the homologous temperature prescribes that the Ag film grows on both substrates in the zone I described by the structure zone model some significant differences are observed on the basis of the supporting substrate. In the normal incidence condition, on the SiO2/Si surface a dense close-packed Ag film exhibiting a smooth surface is obtained, while on the TiO2/Ti surface a more columnar film morphology is formed. In the oblique incidence condition the columnar morphology for the Ag film occurs both on SiO2/Si and TiO2/Ti but a higher porous columnar film is obtained on TiO2/Ti due to the lower Ag diffusivity. These results indicate that the adatoms diffusivity on the substrate as determined by the adatom-surface interaction (in addition to the substrate temperature) strongly determines the final film nanostructure.

  5. Cognitive deficits in a murine model of the eosinophilia-myalgia syndrome: a preliminary report.

    PubMed

    Middaugh, L D; Nussbaum, R; Ludwicka, A; Bolster, M B; Silver, R M

    1996-01-01

    The described study was to determine the effects of chronic exposure to 1,1'-ethylidenebis[L-tryptophan] (EBT), a tryptophan contaminant, on cognitive behavior of female C57BL/6 (C57) mice. EBT (also designated as "peak E" or "peak 97") is one of several compounds that are suspect in the eosinophilia-myalgia syndrome (EMS). Groups of female C57 mice (12/group) were injected IP with saline (SAL), tryptophan (TRY), EBT, or an EBT + tryptophan combination (EBT + TRY) over a 6-week period. Previous experiments established that the dosing conditions produce several characteristics of EMS, including dermal inflammation and fibrosis, increased dermal mast cells, and increased levels of quinolinic acid. The mice exposed to EBT + TRY were abnormal during the solution of a Morris water maze problem. First, they had a shorter latency to locate the submerged platform goal during the initial day of training compared to SAL or TRY mice; secondly, they did not show the systematic reduction in latency to locate the platform goal across days of training noted for SAL or TRY mice. These abnormalities occurred in the absence of altered body weight or gross motor activity during the treatment procedure, or in the animal's swim speeds at the time of testing, 3 days after termination of treatment. The results suggest that prolonged exposure to EBT + TRY can alter the reaction to a stressful environment and can alter cognitive behavior.

  6. The New Film Technologies: Computerized Video-Assisted Film Production.

    ERIC Educational Resources Information Center

    Mott, Donald R.

    Over the past few years, video technology has been used to assist film directors after they have shot a scene, to control costs, and to create special effects, especially computer assisted graphics. At present, a computer based editing system called "Film 5" combines computer technology and video tape with film to save as much as 50% of…

  7. Contributions of Film Introductions and Film Summaries to Learning from Instructional Films.

    ERIC Educational Resources Information Center

    Lathrop, C. W., Jr.; Norford, C. A.

    An exploratory study of the contribution to learning of typical introductory and summarizing sequences in instructional films underlined the need for further experimental work to determine what kinds of introductory and concluding sequences are most useful in promoting learning from films. The first part of the study was concerned with film…

  8. Exploring the Film.

    ERIC Educational Resources Information Center

    Kuhns, William; Stanley, Robert

    The purpose of film study is defined here in the words of D. W. Griffiths: "My goal is above all to make you see." This book is intended to be used as a text in a film study course. It traces the development of films from a scientific curiosity through silent films to modern wide screen productions. A comic strip is used to demonstrate the effect…

  9. Electrochromic nanocomposite films

    DOEpatents

    Milliron, Delia; Llordes, Anna; Buonsanti, Raffaella; Garcia, Guillermo

    2018-04-10

    The present invention provides an electrochromic nanocomposite film. In an exemplary embodiment, the electrochromic nanocomposite film, includes (1) a solid matrix of oxide based material and (2) transparent conducting oxide (TCO) nanostructures embedded in the matrix. In a further embodiment, the electrochromic nanocomposite film farther includes a substrate upon which the matrix is deposited. The present invention also provides a method of preparing an electrochromic nanocomposite film.

  10. Assessing mental health clinicians' intentions to adopt evidence-based treatments: reliability and validity testing of the evidence-based treatment intentions scale.

    PubMed

    Williams, Nathaniel J

    2016-05-05

    Intentions play a central role in numerous empirically supported theories of behavior and behavior change and have been identified as a potentially important antecedent to successful evidence-based treatment (EBT) implementation. Despite this, few measures of mental health clinicians' EBT intentions exist and available measures have not been subject to thorough psychometric evaluation or testing. This paper evaluates the psychometric properties of the evidence-based treatment intentions (EBTI) scale, a new measure of mental health clinicians' intentions to adopt EBTs. The study evaluates the reliability and validity of inferences made with the EBTI using multi-method, multi-informant criterion variables collected over 12 months from a sample of 197 mental health clinicians delivering services in 13 mental health agencies. Structural, predictive, and discriminant validity evidence is assessed. Findings support the EBTI's factor structure (χ (2) = 3.96, df = 5, p = .556) and internal consistency reliability (α = .80). Predictive validity evidence was provided by robust and significant associations between EBTI scores and clinicians' observer-reported attendance at a voluntary EBT workshop at a 1-month follow-up (OR = 1.92, p < .05), self-reported EBT adoption at a 12-month follow-up (R (2) = .17, p < .001), and self-reported use of EBTs with clients at a 12-month follow-up (R (2) = .25, p < .001). Discriminant validity evidence was provided by small associations with clinicians' concurrently measured psychological work climate perceptions of functionality (R (2) = .06, p < .05), engagement (R (2) = .06, p < .05), and stress (R (2) = .00, ns). The EBTI is a practical and theoretically grounded measure of mental health clinicians' EBT intentions. Scores on the EBTI provide a basis for valid inferences regarding mental health clinicians' intentions to adopt EBTs. Discussion focuses on research and practice applications.

  11. Effect-based trigger values for in vitro and in vivo bioassays performed on surface water extracts supporting the environmental quality standards (EQS) of the European Water Framework Directive.

    PubMed

    Escher, Beate I; Aїt-Aїssa, Selim; Behnisch, Peter A; Brack, Werner; Brion, François; Brouwer, Abraham; Buchinger, Sebastian; Crawford, Sarah E; Du Pasquier, David; Hamers, Timo; Hettwer, Karina; Hilscherová, Klára; Hollert, Henner; Kase, Robert; Kienle, Cornelia; Tindall, Andrew J; Tuerk, Jochen; van der Oost, Ron; Vermeirssen, Etienne; Neale, Peta A

    2018-07-01

    Effect-based methods including cell-based bioassays, reporter gene assays and whole-organism assays have been applied for decades in water quality monitoring and testing of enriched solid-phase extracts. There is no common EU-wide agreement on what level of bioassay response in water extracts is acceptable. At present, bioassay results are only benchmarked against each other but not against a consented measure of chemical water quality. The EU environmental quality standards (EQS) differentiate between acceptable and unacceptable surface water concentrations for individual chemicals but cannot capture the thousands of chemicals in water and their biological action as mixtures. We developed a method that reads across from existing EQS and includes additional mixture considerations with the goal that the derived effect-based trigger values (EBT) indicate acceptable risk for complex mixtures as they occur in surface water. Advantages and limitations of various approaches to read across from EQS are discussed and distilled to an algorithm that translates EQS into their corresponding bioanalytical equivalent concentrations (BEQ). The proposed EBT derivation method was applied to 48 in vitro bioassays with 32 of them having sufficient information to yield preliminary EBTs. To assess the practicability and robustness of the proposed approach, we compared the tentative EBTs with observed environmental effects. The proposed method only gives guidance on how to derive EBTs but does not propose final EBTs for implementation. The EBTs for some bioassays such as those for estrogenicity are already mature and could be implemented into regulation in the near future, while for others it will still take a few iterations until we can be confident of the power of the proposed EBTs to differentiate good from poor water quality with respect to chemical contamination. Copyright © 2018 Elsevier B.V. All rights reserved.

  12. The American Film Heritage; Impressions from the American Film Institute Archives.

    ERIC Educational Resources Information Center

    Shales, Tom; And Others

    The American Film Institute has an archive which presently contains more than 9,000 films, many of them rare. The articles in this volume are based on some of the films in the collection. Among the topics of these essays are: pioneers like D. W. Griffith and Thomas H. Ince, treatment of blacks and Indians in films, development of the techniques…

  13. International Film Guide 1972.

    ERIC Educational Resources Information Center

    Cowie, Peter, Ed.

    Brief descriptions of film production activity in 44 countries and reviews of important films from these countries are presented, along with lists of films recently in production. A paragraph-long description of film festivals slated for 1972 is given. Sound track music available on records, 16mm films available in the United Kingdom and the…

  14. Science Fiction on Film.

    ERIC Educational Resources Information Center

    Burmester, David

    1985-01-01

    Reviews science fiction films used in a science fiction class. Discusses feature films, short science fiction films, short story adaptations, original science fiction pieces and factual science films that enrich literature. (EL)

  15. Lights … action: Comparison of trauma films for use in the trauma film paradigm.

    PubMed

    Arnaudova, Inna; Hagenaars, Muriel A

    2017-06-01

    Affective films are often used in emotion research and negative films are frequently used as an analogue for trauma (trauma film paradigm). However, different films are used with possibly distinct consequences. We aimed to investigate specific effects of four negative films covering distinct themes (physical, sexual, traffic and food), and tested neutral and positive films with matching content. Self-reported emotional responses and heart rate during the films were examined (immediate responses) as well as intrusions of the films in the subsequent week (delayed responses). Within each theme, negative films were rated as more unpleasant than the positive and neutral counterparts. They also evoked more negative emotions and more intrusive memories. Across themes, the four negative films did not differ in terms of valence and arousal, but clearly differed on immediate (e.g., disgust, embarrassment, heart rate) and delayed (intrusions) effects. Thus, we urge researchers to carefully select negative films for their studies, as different films seem to evoke distinct emotional responses. In addition, using positive films within the same themes is recommended in order to control for effects of arousal. In general, the specific film material should be considered when comparing effects across studies. Copyright © 2017 Elsevier Ltd. All rights reserved.

  16. Thick Film Interference.

    ERIC Educational Resources Information Center

    Trefil, James

    1983-01-01

    Discusses why interference effects cannot be seen with a thick film, starting with a review of the origin of interference patterns in thin films. Considers properties of materials in films, properties of the light source, and the nature of light. (JN)

  17. Electrochemical Properties of Si Film Electrodes Containing TiNi Thin-Film Current Collectors

    NASA Astrophysics Data System (ADS)

    Im, Yeon-min; Noh, Jung-pil; Cho, Gyu-bong; Nam, Tea-hyun

    2018-03-01

    A 50.3Ti-49.7Ni thin film fabricated by DC sputtering was employed as a current collector of Si film electrode. The structural and electrochemical properties of Si/TiNi film electrode were compared with those of a Si/Cu film electrode. The TiNi film with cluster-like structures composed of B2 austenitic phase at room temperature displayed the high electrochemical stability for Li ions. The amorphous Si film deposited on the TiNi film also consisted of cluster-like structures on the surface. The Si film grown on the TiNi film current collector (Si/TiNi electrode) demonstrated a high columbic efficiency of 87% at the first cycle (363 μAh/cm2 of charge capacity and 314 μAh/cm2 of discharge capacity). The Si/TiNi electrode exhibited better electrochemical properties in terms of capacity, cycle performance, and structural stability compared to the Si electrode with a conventional Cu foil current collector.

  18. A Nonlinear Finite Element Framework for Viscoelastic Beams Based on the High-Order Reddy Beam Theory

    DTIC Science & Technology

    2012-06-09

    employed theories are the Euler-Bernoulli beam theory (EBT) and the Timoshenko beam theory ( TBT ). The major deficiency associated with the EBT is failure to...account for defor- mations associated with shearing. The TBT relaxes the normality assumption of the EBT and admits a constant state of shear strain...on a given cross-section. As a result, the TBT necessitates the use of shear correction coefficients in order to accurately predict transverse

  19. Exhaled Breath Temperature as a Novel Marker of Future Development of COPD: Results of a Follow-Up Study in Smokers.

    PubMed

    Labor, Marina; Vrbica, Žarko; Gudelj, Ivan; Labor, Slavica; Jurić, Iva; Plavec, Davor

    2016-12-01

    Although only less than one-third of smokers develop COPD, early marker(s) of COPD development are lacking. The aim of this research was to assess the ability of an average equilibrium exhaled breath temperature (EBT) in identifying susceptibility to cigarette smoke so as to predict COPD development in smokers at risk. The study was a part of a multicenter prospective cohort study in current smokers (N = 140, both sexes, 40-65 years, ≥20 pack-years) with no prior diagnosis of COPD. Diagnostic workup includes history, physical, quality of life, hematology and highly sensitive CRP, EBT before and after smoking a cigarette, lung function with bronchodilator test, and 6-minute walk test. Patients without a diagnosis of COPD and in GOLD 1 stage at initial assessment were reassessed after 2 years. COPD was additionally diagnosed based on lower level of normal (LLN) lung function criteria. Utility of EBT for disease progression was analyzed using receiver operator curve (ROC) and logistic regression analyses. Change in EBT after smoking a cigarette at initial visit (ΔEBT) was significantly predictive for disease progression (newly diagnosed COPD; newly diagnosed COPD + severity progression) after 2 years (p < 0.05 for both). ΔEBT had an AUC of 0.859 (p = 0.011) with sensitivity of 66.7% and specificity of 98.1% for newly diagnosed COPD using LLN criteria. We conclude that EBT shows potential for predicting the future development of COPD in current smokers. This was best seen using LLN to diagnose COPD, adding further evidence to question the use of GOLD criteria for diagnosing COPD.

  20. SU-E-T-82: Comparison of Several Lumbar Intervertebral Fusion Titanium Cages with Respect to Their Backscattering Properties

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Failing, T; Chofor, N; Poppinga, D

    Purpose: Investigating the backscatter dose factor with regards to structure and geometry of the surface material. Methods: The titanium cages used for this study representing both prototypes and well established products are made of a laser-sintered titanium alloy (AditusV GmbH, Berlin, Germany). A set of four radiochromic EBT3 films was used in a stacked geometry to measure the range and the magnitude of the expected surface dose enhancement due to the in comparison to water increased secondary electron release from the material. The measurement geometry and the small thickness of radiochromic EBT3 film allowed the dose measurement at distances ofmore » 0.1 mm, 0.9 mm, 1.7 mm and 2.5 mm from the probe surfaces. Water reference measurements were taken under equal conditions, in order to allow the calculation of the relative dose enhancement at the surface of a probe. Measurements were performed within a water phantom. An Epson Expression 10000 XL flatbed scanner was used for digitization. Results: Sintered titanium showed a dose enhancement factor of 1.22 at the surface of the material. The factor can be reduced to less than 1.10 by utilizing mesh structures. In both cases, the dose enhancement factor decreased to less than 1.03 at a distance of 1.7mm indicating the low energy of scattered electrons. Conclusion: Backscattering of titanium cages should be considered in treatment planning, especially when the cages are located close to organs at risk. While mesh structures were introduced to improve bone fusion with the implant structure, the potentially harmful surface dose enhancement is significantly reduced.« less