Science.gov

Sample records for mammography image quality

  1. Quality Imaging - Comparison of CR Mammography with Screen-Film Mammography

    SciTech Connect

    Gaona, E.; Azorin Nieto, J.; Iran Diaz Gongora, J. A.; Arreola, M.; Casian Castellanos, G.; Perdigon Castaneda, G. M.; Franco Enriquez, J. G.

    2006-09-08

    The aim of this work is a quality imaging comparison of CR mammography images printed to film by a laser printer with screen-film mammography. A Giotto and Elscintec dedicated mammography units with fully automatic exposure and a nominal large focal spot size of 0.3 mm were used for the image acquisition of phantoms in screen-film mammography. Four CR mammography units from two different manufacturers and three dedicated x-ray mammography units with fully automatic exposure and a nominal large focal spot size of 0.3 mm were used for the image acquisition of phantoms in CR mammography. The tests quality image included an assessment of system resolution, scoring phantom images, Artifacts, mean optical density and density difference (contrast). In this study, screen-film mammography with a quality control program offers a significantly greater level of quality image relative to CR mammography images printed on film.

  2. TL dosimetry for quality control of CR mammography imaging systems

    NASA Astrophysics Data System (ADS)

    Gaona, E.; Nieto, J. A.; Góngora, J. A. I. D.; Arreola, M.; Enríquez, J. G. F.

    The aim of this work is to estimate the average glandular dose with thermoluminescent (TL) dosimetry and comparison with quality imaging in computed radiography (CR) mammography. For a measuring dose, the Food and Drug Administration (FDA) and the American College of Radiology (ACR) use a phantom, so that dose and image quality are assessed with the same test object. The mammography is a radiological image to visualize early biological manifestations of breast cancer. Digital systems have two types of image-capturing devices, full field digital mammography (FFDM) and CR mammography. In Mexico, there are several CR mammography systems in clinical use, but only one system has been approved for use by the FDA. Mammography CR uses a photostimulable phosphor detector (PSP) system. Most CR plates are made of 85% BaFBr and 15% BaFI doped with europium (Eu) commonly called barium flourohalideE We carry out an exploratory survey of six CR mammography units from three different manufacturers and six dedicated X-ray mammography units with fully automatic exposure. The results show three CR mammography units (50%) have a dose greater than 3.0 mGy without demonstrating improved image quality. The differences between doses averages from TLD system and dosimeter with ionization chamber are less than 10%. TLD system is a good option for average glandular dose measurement for X-rays with a HVL (0.35-0.38 mmAl) and kVp (24-26) used in quality control procedures with ACR Mammography Accreditation Phantom.

  3. The influence of software filtering in digital mammography image quality

    NASA Astrophysics Data System (ADS)

    Michail, C.; Spyropoulou, V.; Kalyvas, N.; Valais, I.; Dimitropoulos, N.; Fountos, G.; Kandarakis, I.; Panayiotakis, G.

    2009-05-01

    Breast cancer is one of the most frequently diagnosed cancers among women. Several techniques have been developed to help in the early detection of breast cancer such as conventional and digital x-ray mammography, positron and single-photon emission mammography, etc. A key advantage in digital mammography is that images can be manipulated as simple computer image files. Thus non-dedicated commercially available image manipulation software can be employed to process and store the images. The image processing tools of the Photoshop (CS 2) software usually incorporate digital filters which may be used to reduce image noise, enhance contrast and increase spatial resolution. However, improving an image quality parameter may result in degradation of another. The aim of this work was to investigate the influence of three sharpening filters, named hereafter sharpen, sharpen more and sharpen edges on image resolution and noise. Image resolution was assessed by means of the Modulation Transfer Function (MTF).In conclusion it was found that the correct use of commercial non-dedicated software on digital mammograms may improve some aspects of image quality.

  4. Average glandular dose and phantom image quality in mammography

    NASA Astrophysics Data System (ADS)

    Oliveira, M.; Nogueira, M. S.; Guedes, E.; Andrade, M. C.; Peixoto, J. E.; Joana, G. S.; Castro, J. G.

    2007-09-01

    Doses in mammography should be maintained as low as possible without reducing the high image quality needed for early detection of the breast cancer. The breast is composed of tissues with very close composition and densities. It increases the difficulty to detect small changes in the normal anatomical structures which may be associated with breast cancer. To achieve the standards of definition and contrast for mammography, the quality and intensity of the X-ray beam, the breast positioning and compression, the film-screen system, and the film processing have to be in optimal operational conditions. This study sought to evaluate average glandular dose (AGD) and image quality on a standard phantom in 134 mammography units in the state of Minas Gerais, Brazil, between December 2004 and May 2006. AGDs were obtained by means of entrance kerma measured with TL LiF100 dosimeters on phantom surface. Phantom images were obtained with automatic exposure technique, fixed 28 kV and molybdenum anode-filter combination. The phantom used contained structures simulating tumoral masses, microcalcifications, fibers and low contrast areas. High-resolution metallic meshes to assess image definition and a stepwedge to measure image contrast index were also inserted in the phantom. The visualization of simulated structures, the mean optical density and the contrast index allowed to classify the phantom image quality in a seven-point scale. The results showed that 54.5% of the facilities did not achieve the minimum performance level for image quality. It is mainly due to insufficient film processing observed in 61.2% of the units. AGD varied from 0.41 to 2.73 mGy with a mean value of 1.32±0.44 mGy. In all optimal quality phantom images, AGDs were in this range. Additionally, in 7.3% of the mammography units, the AGD constraint of 2 mGy was exceeded. One may conclude that dose level to patient and image quality are not in conformity to regulations in most of the facilities. This indicates that ongoing actions are needed to optimize image quality and radiation dose for early detection of the breast cancer.

  5. Effect of scatter on image quality in synchrotron radiation mammography

    NASA Astrophysics Data System (ADS)

    Moeckli, Raphael; Verdun, Francis R.; Fiedler, Stefan; Pachoud, Marc; Schnyder, Pierre; Valley, Jean-Francois

    2001-06-01

    The display of low-contrast structures and fine microcalcifications is essential for the early diagnosis of breast cancer. In order to achieve a high image quality level with a minimum amount of radiation delivered to the patient, the use of different spectra (Mo or Rh anode and filters) was introduced. The European Synchrotron Radiation Facility is able to produce a monochromatic beam with a high photon flux. It is thus a powerful tool to study the effect of beam energy on image quality and dose in mammography. Our image quality assessment is based on the calculation of the size of the smallest microcalcification detectable on a radiograph, derived from the statistical decision theory. The mean glandular dose is simultaneously measured. Compared with conventional mammography units, the monochromaticity of synchrotron beams improves contrast and the use of a slit instead of an anti-scatter grid leads to a higher primary beam transmission. The relative contribution of these two effects on image quality and dose is discussed.

  6. Comparison of Image Quality Criteria between Digital Storage Phosphor Plate in Mammography and Full-Field Digital Mammography in the Detection of Breast Cancer

    PubMed Central

    Thevi Rajendran, Pushpa; Krishnapillai, Vijayalakshmi; Tamanang, Sulaiman; Kumari Chelliah, Kanaga

    2012-01-01

    Background: Digital mammography is slowly replacing screen film mammography. In digital mammography, 2 methods are available in acquiring images: digital storage phosphor plate and full-field digital mammography. The aim of this study was to compare the image quality acquired from the 2 methods of digital mammography in the detection of breast cancer. Methods: The study took place at the National Cancer Society, Kuala Lumpur, and followed 150 asymptomatic women for the duration of 1 year. Participating women gave informed consent and were exposed to 4 views from each system. Two radiologists independently evaluated the printed images based on the image quality criteria in mammography. McNemar’s test was used to compare the image quality criteria between the systems. Results: The agreement between the radiologists for the digital storage phosphor plate was ? = 0.551 and for full-field digital mammography was ? = 0.523. Full-field digital mammography was significantly better compared with the digital storage phosphor plate in right and left mediolateral oblique views (P < 0.05) in the detection of microcalcifications, which are early signs of breast cancer. However, both systems were comparable in all other aspects of image quality. Conclusion: Digital mammography is a useful screening tool for the detection of early breast cancer and ensures better prognosis and quality of life. PMID:22977375

  7. Mammography (image)

    MedlinePLUS

    Mammography is a low-powered x-ray technique that captures a picture of the internal structure of ... magnified views are taken of suspicious areas. A mammogram may help in the diagnosis of breast problems, ...

  8. Effect of image quality on calcification detection in digital mammography

    PubMed Central

    Warren, Lucy M.; Mackenzie, Alistair; Cooke, Julie; Given-Wilson, Rosalind M.; Wallis, Matthew G.; Chakraborty, Dev P.; Dance, David R.; Bosmans, Hilde; Young, Kenneth C.

    2012-01-01

    Purpose: This study aims to investigate if microcalcification detection varies significantly when mammographic images are acquired using different image qualities, including: different detectors, dose levels, and different image processing algorithms. An additional aim was to determine how the standard European method of measuring image quality using threshold gold thickness measured with a CDMAM phantom and the associated limits in current EU guidelines relate to calcification detection. Methods: One hundred and sixty two normal breast images were acquired on an amorphous selenium direct digital (DR) system. Microcalcification clusters extracted from magnified images of slices of mastectomies were electronically inserted into half of the images. The calcification clusters had a subtle appearance. All images were adjusted using a validated mathematical method to simulate the appearance of images from a computed radiography (CR) imaging system at the same dose, from both systems at half this dose, and from the DR system at quarter this dose. The original 162 images were processed with both Hologic and Agfa (Musica-2) image processing. All other image qualities were processed with Agfa (Musica-2) image processing only. Seven experienced observers marked and rated any identified suspicious regions. Free response operating characteristic (FROC) and ROC analyses were performed on the data. The lesion sensitivity at a nonlesion localization fraction (NLF) of 0.1 was also calculated. Images of the CDMAM mammographic test phantom were acquired using the automatic setting on the DR system. These images were modified to the additional image qualities used in the observer study. The images were analyzed using automated software. In order to assess the relationship between threshold gold thickness and calcification detection a power law was fitted to the data. Results: There was a significant reduction in calcification detection using CR compared with DR: the alternative FROC (AFROC) area decreased from 0.84 to 0.63 and the ROC area decreased from 0.91 to 0.79 (p < 0.0001). This corresponded to a 30% drop in lesion sensitivity at a NLF equal to 0.1. Detection was also sensitive to the dose used. There was no significant difference in detection between the two image processing algorithms used (p > 0.05). It was additionally found that lower threshold gold thickness from CDMAM analysis implied better cluster detection. The measured threshold gold thickness passed the acceptable limit set in the EU standards for all image qualities except half dose CR. However, calcification detection varied significantly between image qualities. This suggests that the current EU guidelines may need revising. Conclusions: Microcalcification detection was found to be sensitive to detector and dose used. Standard measurements of image quality were a good predictor of microcalcification cluster detection. PMID:22755704

  9. Effect of image quality on calcification detection in digital mammography

    SciTech Connect

    Warren, Lucy M.; Mackenzie, Alistair; Cooke, Julie; Given-Wilson, Rosalind M.; Wallis, Matthew G.; Chakraborty, Dev P.; Dance, David R.; Bosmans, Hilde; Young, Kenneth C.

    2012-06-15

    Purpose: This study aims to investigate if microcalcification detection varies significantly when mammographic images are acquired using different image qualities, including: different detectors, dose levels, and different image processing algorithms. An additional aim was to determine how the standard European method of measuring image quality using threshold gold thickness measured with a CDMAM phantom and the associated limits in current EU guidelines relate to calcification detection. Methods: One hundred and sixty two normal breast images were acquired on an amorphous selenium direct digital (DR) system. Microcalcification clusters extracted from magnified images of slices of mastectomies were electronically inserted into half of the images. The calcification clusters had a subtle appearance. All images were adjusted using a validated mathematical method to simulate the appearance of images from a computed radiography (CR) imaging system at the same dose, from both systems at half this dose, and from the DR system at quarter this dose. The original 162 images were processed with both Hologic and Agfa (Musica-2) image processing. All other image qualities were processed with Agfa (Musica-2) image processing only. Seven experienced observers marked and rated any identified suspicious regions. Free response operating characteristic (FROC) and ROC analyses were performed on the data. The lesion sensitivity at a nonlesion localization fraction (NLF) of 0.1 was also calculated. Images of the CDMAM mammographic test phantom were acquired using the automatic setting on the DR system. These images were modified to the additional image qualities used in the observer study. The images were analyzed using automated software. In order to assess the relationship between threshold gold thickness and calcification detection a power law was fitted to the data. Results: There was a significant reduction in calcification detection using CR compared with DR: the alternative FROC (AFROC) area decreased from 0.84 to 0.63 and the ROC area decreased from 0.91 to 0.79 (p < 0.0001). This corresponded to a 30% drop in lesion sensitivity at a NLF equal to 0.1. Detection was also sensitive to the dose used. There was no significant difference in detection between the two image processing algorithms used (p > 0.05). It was additionally found that lower threshold gold thickness from CDMAM analysis implied better cluster detection. The measured threshold gold thickness passed the acceptable limit set in the EU standards for all image qualities except half dose CR. However, calcification detection varied significantly between image qualities. This suggests that the current EU guidelines may need revising. Conclusions: Microcalcification detection was found to be sensitive to detector and dose used. Standard measurements of image quality were a good predictor of microcalcification cluster detection.

  10. Investigation of diagnostic and image quality attributes: comparison of screen-film to CR mammography

    NASA Astrophysics Data System (ADS)

    Fletcher-Heath, Lynn; Richards, Anne; Ryan-Kron, Susan

    2006-03-01

    Digital mammography is advancing into an arena where analog has long been the gold standard. Direct digital systems may not be the favored solution for a particular site while computed radiography (CR) mammography, remains unproven worldwide. This pilot study responds to the growing desire to acquire and display digital mammographic images by exploring the acceptability of CR mammography. Images representing a range of breast tissue types were collected from 49 subjects (17 screening; 32 diagnostic) at four clinical sites. Comparison views were collected on the same breast, under the same compression, using automatic exposure control on state-of-the-art film systems followed by CR. CR images were processed and printed to a mammography printer for hard copy feature comparison. Each image pair in the study was evaluated according to 13 image quality attributes covering noise, contrast, sharpness, and image quality in the overall captured images as well as in each of several particular breast regions (periphery and skin-line, parenchyma and fatty tissue). A rating scale from 1 to 5 was used (strong preference for film=1, strong preference for CR=5). Twelve experienced mammographers at four clinical sites scored a subset of the 49 cases for a total of 64 image pair readings. There were 64 ratings for each of 13 image quality attributes for all cases and, an additional series of scores (four or five attribute ratings) for each abnormality in the category of mass, architectural distortion and microcalcification, for a total of 1069 scores. Based on the pilot study results, it was suggested that CR was equivalent or preferred to conventional screen-film for overall image quality (38% scored 3; 46% scored >3), image contrast (27% scored 3; 59% scored >3) and sharpness (28% scored 3; 50% scored >3). No preference was found when assessing noise. This pilot study also suggested that diagnostic quality was maintained in assessing abnormalities for attributes necessary to evaluate masses and microcalcifications as compared to screen-film.

  11. Figure of Image Quality and Information Capacity in Digital Mammography

    PubMed Central

    Michail, Christos M.; Kalyvas, Nektarios E.; Valais, Ioannis G.; Fudos, Ioannis P.; Fountos, George P.; Dimitropoulos, Nikos; Kandarakis, Ioannis S.

    2014-01-01

    Objectives. In this work, a simple technique to assess the image quality characteristics of the postprocessed image is developed and an easy to use figure of image quality (FIQ) is introduced. This FIQ characterizes images in terms of resolution and noise. In addition information capacity, defined within the context of Shannon's information theory, was used as an overall image quality index. Materials and Methods. A digital mammographic image was postprocessed with three digital filters. Resolution and noise were calculated via the Modulation Transfer Function (MTF), the coefficient of variation, and the figure of image quality. In addition, frequency dependent parameters such as the noise power spectrum (NPS) and noise equivalent quanta (NEQ) were estimated and used to assess information capacity. Results. FIQs for the “raw image” data and the image processed with the “sharpen edges” filter were found 907.3 and 1906.1, correspondingly. The information capacity values were 60.86 × 103 and 78.96 × 103?bits/mm2. Conclusion. It was found that, after the application of the postprocessing techniques (even commercial nondedicated software) on the raw digital mammograms, MTF, NPS, and NEQ are improved for medium to high spatial frequencies leading to resolving smaller structures in the final image. PMID:24895593

  12. SU-E-I-04: A Mammography Phantom to Measure Mean Glandular Dose and Image Quality

    SciTech Connect

    Lopez-Pineda, E; Ruiz-Trejo, C; E, Brandan M

    2014-06-01

    Purpose: To evaluate mean glandular dose (MGD) and image quality in a selection of mammography systems using a novel phantom based on thermoluminescent dosemeters and the ACR wax insert. Methods: The phantom consists of two acrylic, 19 cm diameter, 4.5 cm thick, semicircular modules, used in sequence. The image quality module contains the ACR insert and is used to obtain a quality control image under automatic exposure conditions. The dosimetric module carries 15 TLD-100 chips, some under Al foils, to determine air kerma and half-value-layer. TL readings take place at our laboratory under controlled conditions. Calibration was performed using an ionization chamber and a Senographe 2000D unit for a variety of beam qualities, from 24 to 40 kV, Mo and Rh anodes and filters. Phantom MGD values agree, on the average, within 3% with ionization chamber data, and their precision is better than 10% (k=1). Results: MGD and image quality have been evaluated in a selection of mammography units currently used in Mexican health services. The sample includes analogic (screen/film), flexible digital (CR), and full-field digital image receptors. The highest MDG are associated to the CR technology. The most common image quality failure is due to artifacts (dust, intensifying screen scratches, and processor marks for film/screen, laser reader defects for CR). Conclusion: The developed phantom permits the MGD measurement without the need of a calibrated ionization chamber at the mammography site and can be used by a technician without the presence of a medical physicist. The results indicate the urgent need to establish quality control programs for mammography.

  13. Mammography in New Zealand: radiation dose and image quality.

    PubMed

    Poletti, J L; Williamson, B D; Mitchell, A W

    1991-06-01

    The mean glandular doses to the breast, image quality and machine performance have been determined for all mammographic x-ray facilities in New Zealand, during 1988-89. For 30 mm and 45 mm phantoms the mean doses per film were 1.03 +/- 0.56 mGy and 1.97 +/- 1.06 mGy. These doses are within international guide-lines. Image quality (detection of simulated microcalcifications, and contrast-detail performance) was found to depend on focal spot size/FFD combination, breast thickness, and film processing. The best machines could resolve 0.2 mm aluminium oxide specks with the contact technique. The use of a grid improved image quality as did magnification. Extended cycle film processing reduced doses, but the claimed improvement in image quality was not apparent from our data. The machine calibration parameters kVp, HVL and timer accuracy were in general within accepted tolerances. Automatic exposure controls in some cases gave poor control of film density with changing breast thickness. PMID:1747087

  14. Dosimetry and image quality in digital mammography facilities in the State of Minas Gerais, Brazil

    NASA Astrophysics Data System (ADS)

    da Silva, Sabrina Donato; Joana, Geórgia Santos; Oliveira, Bruno Beraldo; de Oliveira, Marcio Alves; Leyton, Fernando; Nogueira, Maria do Socorro

    2015-11-01

    According to the National Register of Health Care Facilities (CNES), there are approximately 477 mammography systems operating in the state of Minas Gerais, Brazil, of which an estimated 200 are digital apparatus using mainly computerized radiography (CR) or direct radiography (DR) systems. Mammography is irreplaceable in the diagnosis and early detection of breast cancer, the leading cause of cancer death among women worldwide. A high standard of image quality alongside smaller doses and optimization of procedures are essential if early detection is to occur. This study aimed to determine dosimetry and image quality in 68 mammography services in Minas Gerais using CR or DR systems. The data of this study were collected between the years of 2011 and 2013. The contrast-to-noise ratio proved to be a critical point in the image production chain in digital systems, since 90% of services were not compliant in this regard, mainly for larger PMMA thicknesses (60 and 70 mm). Regarding the image noise, only 31% of these were compliant. The average glandular dose found is of concern, since more than half of the services presented doses above acceptable limits. Therefore, despite the potential benefits of using CR and DR systems, the employment of this technology has to be revised and optimized to achieve better quality image and reduce radiation dose as much as possible.

  15. Quality assurance in screening mammography.

    PubMed

    1990-01-01

    This evaluation was a cooperative effort between ECRI and the Swedish Testing Institute for Medical Supplies (Sprima AB), Stockholm. As noted in our last mammography evaluation, we are initially focusing on mammography units in our series of radiology and imaging studies because (1) improvements in technology (e.g., lower radiation doses) now make routine screening feasible; (2) third-party payers are supporting screening programs, and (3) breast cancer is a major threat to women's health. Again ECRI will provide physicists and radiologists with our raw test data on request. In our previous evaluation, we discussed the role of mammography in reducing breast cancer mortality rates and tested the units for screen-film mammography and xeromammography capability, basing our ratings primarily on the units' ability to safely and consistently produce acceptable images with minimal patient dose during screen-film mammography. Below, we focus on the definition, measurement, and significance of image quality within the context of screen-film mammography. We evaluated seven mammography units from seven manufacturers. We also retested the Automatic Exposure Control (AEC) mode of the Soredex Mamex dc Mag, which we evaluated last year and rated Conditionally Acceptable because of its AEC performance (see "Soredex Mamex dc Mag"). Our ratings are based on our minimum criteria of acceptability: that the units safely and reliably produce the best possible image quality while delivering the lowest possible dose to the patient. Except for the Kramex HF-45, all of the evaluated units-the Acoma ESP-200, GE-CGR Senographe 600T, Instrumentarium/Ausonics (I/A) Alpha III, Lorad M II-D, Philips MammoDiagnost UC, Picker Sureview, and the Soredex Mamex dc Mag-exceed our minimum criteria and are rated Acceptable for both screening and magnification (diagnostic) applications. The Kramex HF-45, the only dedicated screening unit we evaluated, is rated Acceptable-Not Recommended because of its limitations in achieving optimal image quality. The units are ranked according to their AEC performance and human factors (ergonomic) design, as well as on technical phantom image quality results, within the two applications. Clinical phantom image quality results are listed, but were not used to rank the units. Refer to our previous study for further discussions of breast imaging technology and terminology, as well as radiation risks. Terms set below in small caps are defined in the Glossary. PMID:2372321

  16. A new test phantom with different breast tissue compositions for image quality assessment in conventional and digital mammography

    NASA Astrophysics Data System (ADS)

    Pachoud, Marc; Lepori, D.; Valley, Jean-François; Verdun, Francis R.

    2004-12-01

    Our objective is to describe a new test phantom that permits the objective assessment of image quality in conventional and digital mammography for different types of breast tissue. A test phantom, designed to represent a compressed breast, was made from tissue equivalent materials. Three separate regions, with different breast tissue compositions, are used to evaluate low and high contrast resolution, spatial resolution and image noise. The phantom was imaged over a range of kV using a Contour 2000 (Bennett) mammography unit with a Kodak MinR 2190-MinR L screen film combination and a Senograph 2000D (General Electric) digital mammography unit. Objective image quality assessments for different breast tissue compositions were performed using the phantom for conventional and digital mammography. For a similar mean glandular dose (MGD), the digital system gives a significantly higher contrast-to-noise ratio (CNR) than the screen film system for 100% glandular tissue. In conclusion, in mammography, a range of exposure conditions is used for imaging because of the different breast tissue compositions encountered clinically. Ideally, the patient dose image quality relationship should be optimized over the range of exposure conditions. The test phantom presented in this work permits image quality parameters to be evaluated objectively for three different types of breast tissue. Thus, it is a useful tool for optimizing the patient dose image quality relationship.

  17. Evaluation of the quality of image for various breast composition and exposure conditions in digital mammography

    NASA Astrophysics Data System (ADS)

    Yamada, Maki; Kato, Yuri; Fujita, Naotoshi; Kodera, Yoshie

    2011-03-01

    Breast density has a close relationship with breast cancer risk. The exposure parameters must be appropriately chosen for each breast. However, the optimal exposure conditions for digital mammography are uncertain in clinical. The exposure parameters in digital mammography must be optimized with maximization of image quality and minimization of radiation dose. We evaluated image quality under different exposure conditions to investigate the most advantageous tube voltage. For different compressed breast phantom thicknesses and compositions, we measured the Wiener spectrum (WS), noise-equivalent number of quanta (NEQ), and detective quantum efficiency (DQE). In this study, the signal-to-noise ratios were derived from a perceived statistical decision theory model with the internal noise of eye-brain system (SNRi), contrived and studied by Loo et al.1 and Ishida et al.2 These were calculated under a fixed average glandular dose. The WS values were obtained with a fixed image contrast. For 4-cm-thick and 50% glandular breast phantoms, the NEQ showed that high voltages gave a superior noise property of images, especially for thick breasts, but the improvement in the NEQ by tube voltage was not so remarkable. On the other hand, the SNRi value with a Mo filter was larger than that with a Rh filter. The SNRi increased when the tube voltage decreased. The result differed from those of WS and NEQ. In this study, the SNRi depended on the contrast of signal. Accuracy should be high with an intense, low-contrast object.

  18. A phantom using titanium and Landolt rings for image quality evaluation in mammography

    NASA Astrophysics Data System (ADS)

    de las Heras, Hugo; Schöfer, Felix; Tiller, Britta; Chevalier, Margarita; Zwettler, Georg; Semturs, Friedrich

    2013-04-01

    A phantom for image quality evaluation of digital mammography systems is presented and compared to the most widely used phantoms in Europe and the US. The phantom contains objects for subjective detection of Landolt rings (four-alternative, forced-choice task) and for objective calculation of signal-difference-to-noise ratios (SDNR), both in a titanium background within a 12-step wedge. Evaluating phantom images corresponding to exposures between 15 and 160 mAs (average glandular dose between 0.2 and 2 mGy), the resulting scores were compared to the scores obtained following the European EPQC and American College of Radiology (ACR) protocols. Scores of the Landolt test equal to 19 and 8.5 and SDNR equal to 20 and 11 were found to be equivalent to the acceptable limiting values suggested by EPQC and ACR. In addition, the Landolt and SDNR tests were shown to take into account the anatomical variations in thickness and tissue density within the breast. The simplified evaluation method presented was shown to be a sensitive, efficient and reliable alternative for image quality evaluation of mammography systems.

  19. On image quality metrics and the usefulness of grids in digital mammography

    PubMed Central

    Chen, Han; Danielsson, Mats; Xu, Cheng; Cederström, Björn

    2015-01-01

    Abstract. Antiscatter grids are used in digital mammography to reduce the scattered radiation from the breast and improve image contrast. They are, however, imperfect and lead to partial absorption of primary radiation, as well as failing to absorb all scattered radiation. Nevertheless, the general consensus has been that antiscatter grids improve image quality for the majority of breast types and sizes. There is, however, inconsistency in the literature, and recent results show that a substantial image quality improvement can be achieved even for thick breasts if the grid is disposed of. The purpose of this study was to investigate if differences in the considered imaging task and experimental setup could explain the different outcomes. We estimated the dose reduction that can be achieved if the grid were to be removed as a function of breast thickness with varying geometries and experimental conditions. Image quality was quantified by the signal-difference-to-noise ratio (SDNR) measured using an aluminum (Al) filter on blocks of poly(methyl methacrylate) (PMMA), and images were acquired with and without grid at a constant exposure. We also used a theoretical model validated with Monte Carlo simulations. Both theoretically and experimentally, the main finding was that when a large 4×8  cm2 Al filter was used, the SDNR values for the gridless images were overestimated up to 25% compared to the values for the small 1×1  cm2 filter, and gridless imaging was superior for any PMMA thickness. For the small Al filter, gridless imaging was only superior for PMMAs thinner than 4 cm. This discrepancy can be explained by a different sensitivity to and sampling of the angular scatter spread function, depending on the size of the contrast object. The experimental differences were eliminated either by using a smaller region of interest close to the edge of the large filter or by applying a technique of scatter correction by subtracting the estimated scatter image. These results explain the different conclusions reported in the literature and show the importance of the selection of measurement methods. Since the interesting structures in mammography are below the 1-cm scale, we advocate the use of smaller contrast objects for assessment of antiscatter grid performance. PMID:26158077

  20. Image quality, threshold contrast and mean glandular dose in CR mammography

    NASA Astrophysics Data System (ADS)

    Jakubiak, R. R.; Gamba, H. R.; Neves, E. B.; Peixoto, J. E.

    2013-09-01

    In many countries, computed radiography (CR) systems represent the majority of equipment used in digital mammography. This study presents a method for optimizing image quality and dose in CR mammography of patients with breast thicknesses between 45 and 75 mm. Initially, clinical images of 67 patients (group 1) were analyzed by three experienced radiologists, reporting about anatomical structures, noise and contrast in low and high pixel value areas, and image sharpness and contrast. Exposure parameters (kV, mAs and target/filter combination) used in the examinations of these patients were reproduced to determine the contrast-to-noise ratio (CNR) and mean glandular dose (MGD). The parameters were also used to radiograph a CDMAM (version 3.4) phantom (Artinis Medical Systems, The Netherlands) for image threshold contrast evaluation. After that, different breast thicknesses were simulated with polymethylmethacrylate layers and various sets of exposure parameters were used in order to determine optimal radiographic parameters. For each simulated breast thickness, optimal beam quality was defined as giving a target CNR to reach the threshold contrast of CDMAM images for acceptable MGD. These results were used for adjustments in the automatic exposure control (AEC) by the maintenance team. Using optimized exposure parameters, clinical images of 63 patients (group 2) were evaluated as described above. Threshold contrast, CNR and MGD for such exposure parameters were also determined. Results showed that the proposed optimization method was effective for all breast thicknesses studied in phantoms. The best result was found for breasts of 75 mm. While in group 1 there was no detection of the 0.1 mm critical diameter detail with threshold contrast below 23%, after the optimization, detection occurred in 47.6% of the images. There was also an average MGD reduction of 7.5%. The clinical image quality criteria were attended in 91.7% for all breast thicknesses evaluated in both patient groups. Finally, this study also concluded that the use of the AEC of the x-ray unit based on the constant dose to the detector may bring some difficulties to CR systems to operate under optimal conditions. More studies must be performed, so that the compatibility between systems and optimization methodologies can be evaluated, as well as this optimization method. Most methods are developed for phantoms, so comparative studies including clinical images must be developed.

  1. Computer analysis of mammography phantom images (CAMPI)

    NASA Astrophysics Data System (ADS)

    Chakraborty, Dev P.

    1997-05-01

    Computer analysis of mammography phantom images (CAMPI) is a method for objective and precise measurements of phantom image quality in mammography. This investigation applied CAMPI methodology to the Fischer Mammotest Stereotactic Digital Biopsy machine. Images of an American College of Radiology phantom centered on the largest two microcalcification groups were obtained on this machine under a variety of x-ray conditions. Analyses of the images revealed that the precise behavior of the CAMPI measures could be understood from basic imaging physics principles. We conclude that CAMPI is sensitive to subtle image quality changes and can perform accurate evaluations of images, especially of directly acquired digital images.

  2. Positron emission mammography imaging

    SciTech Connect

    Moses, William W.

    2003-10-02

    This paper examines current trends in Positron Emission Mammography (PEM) instrumentation and the performance tradeoffs inherent in them. The most common geometry is a pair of parallel planes of detector modules. They subtend a larger solid angle around the breast than conventional PET cameras, and so have both higher efficiency and lower cost. Extensions to this geometry include encircling the breast, measuring the depth of interaction (DOI), and dual-modality imaging (PEM and x-ray mammography, as well as PEM and x-ray guided biopsy). The ultimate utility of PEM may not be decided by instrument performance, but by biological and medical factors, such as the patient to patient variation in radiotracer uptake or the as yet undetermined role of PEM in breast cancer diagnosis and treatment.

  3. A prototype of mammography CADx scheme integrated to imaging quality evaluation techniques

    NASA Astrophysics Data System (ADS)

    Schiabel, Homero; Matheus, Bruno R. N.; Angelo, Michele F.; Patrocínio, Ana Claudia; Ventura, Liliane

    2011-03-01

    As all women over the age of 40 are recommended to perform mammographic exams every two years, the demands on radiologists to evaluate mammographic images in short periods of time has increased considerably. As a tool to improve quality and accelerate analysis CADe/Dx (computer-aided detection/diagnosis) schemes have been investigated, but very few complete CADe/Dx schemes have been developed and most are restricted to detection and not diagnosis. The existent ones usually are associated to specific mammographic equipment (usually DR), which makes them very expensive. So this paper describes a prototype of a complete mammography CADx scheme developed by our research group integrated to an imaging quality evaluation process. The basic structure consists of pre-processing modules based on image acquisition and digitization procedures (FFDM, CR or film + scanner), a segmentation tool to detect clustered microcalcifications and suspect masses and a classification scheme, which evaluates as the presence of microcalcifications clusters as well as possible malignant masses based on their contour. The aim is to provide enough information not only on the detected structures but also a pre-report with a BI-RADS classification. At this time the system is still lacking an interface integrating all the modules. Despite this, it is functional as a prototype for clinical practice testing, with results comparable to others reported in literature.

  4. Influence of scatter reduction method and monochromatic beams on image quality and dose in mammography.

    PubMed

    Moeckli, Raphaël; Verdun, Francis R; Fiedler, Stefan; Pachoud, Marc; Bulling, Shelley; Schnyder, Pierre; Valley, Jean-François

    2003-12-01

    In mammography, the image contrast and dose delivered to the patient are determined by the x-ray spectrum and the scatter to primary ratio S/P. Thus the quality of the mammographic procedure is highly dependent on the choice of anode and filter material and on the method used to reduce the amount of scattered radiation reaching the detector. Synchrotron radiation is a useful tool to study the effect of beam energy on the optimization of the mammographic process because it delivers a high flux of monochromatic photons. Moreover, because the beam is naturally flat collimated in one direction, a slot can be used instead of a grid for scatter reduction. We have measured the ratio S/P and the transmission factors for grids and slots for monoenergetic synchrotron radiation. In this way the effect of beam energy and scatter rejection method were separated, and their respective importance for image quality and dose analyzed. Our results show that conventional mammographic spectra are not far from optimum and that the use of a slot instead of a grid has an important effect on the optimization of the mammographic process. We propose a simple numerical model to quantify this effect. PMID:14713082

  5. A comparison between objective and subjective image quality measurements for a full field digital mammography system.

    PubMed

    Marshall, N W

    2006-05-21

    This paper presents pre-sampling modulation transfer function (MTF), normalized noise power spectrum (NNPS) and detective quantum efficiency (DQE) results for an amorphous selenium (a-Se) full field digital mammography system. MTF was calculated from the image of an angled 0.5 mm thick Cu edge, acquired without additional beam filtration. NNPS data were acquired at detector air-kerma levels ranging from 9.1 microGy to 331 microGy, using a standard mammography x-ray spectrum of 28 kV, Mo/Mo target/filter combination and 4 cm of PMMA additional filtration. Prior to NNPS estimation, the image statistics were assessed using a variance image. This method was able to easily identify a detector artefact and should prove useful in routine quality assurance (QA) measurements. Detector DQE, calculated from the NNPS and MTF data, dropped to 0.3 for low detector air-kerma settings but reached an approximately constant value of 0.6 above 50 microGy at the detector. Subjective image quality data were also obtained at these detector air-kerma settings using the CDMAM contrast-detail (c-d) test object. The c-d data reflected the trend seen in DQE, with threshold contrast increasing at low detector air-kerma values. The c-d data were then compared against predictions made using two established models, the Rose model and a standard signal detection theory model. Using DQE(0), the Rose model gave results within approximately 15% on average for all the detector air-kerma values studied and for detail diameters down to 0.2 mm. Similar agreement was also found between the measured c-d data and the signal detection theory results, which were calculated using an ideal human visual response function and a system magnification of unity. The use of full spatial frequency DQE improved the agreement between the calculated and observer results for detail sizes below 0.13 mm. PMID:16675862

  6. Image quality assessment in digital mammography: part I. Technical characterization of the systems.

    PubMed

    Marshall, N W; Monnin, P; Bosmans, H; Bochud, F O; Verdun, F R

    2011-07-21

    In many European countries, image quality for digital x-ray systems used in screening mammography is currently specified using a threshold-detail detectability method. This is a two-part study that proposes an alternative method based on calculated detectability for a model observer: the first part of the work presents a characterization of the systems. Eleven digital mammography systems were included in the study; four computed radiography (CR) systems, and a group of seven digital radiography (DR) detectors, composed of three amorphous selenium-based detectors, three caesium iodide scintillator systems and a silicon wafer-based photon counting system. The technical parameters assessed included the system response curve, detector uniformity error, pre-sampling modulation transfer function (MTF), normalized noise power spectrum (NNPS) and detective quantum efficiency (DQE). Approximate quantum noise limited exposure range was examined using a separation of noise sources based upon standard deviation. Noise separation showed that electronic noise was the dominant noise at low detector air kerma for three systems; the remaining systems showed quantum noise limited behaviour between 12.5 and 380 µGy. Greater variation in detector MTF was found for the DR group compared to the CR systems; MTF at 5 mm(-1) varied from 0.08 to 0.23 for the CR detectors against a range of 0.16-0.64 for the DR units. The needle CR detector had a higher MTF, lower NNPS and higher DQE at 5 mm(-1) than the powder CR phosphors. DQE at 5 mm(-1) ranged from 0.02 to 0.20 for the CR systems, while DQE at 5 mm(-1) for the DR group ranged from 0.04 to 0.41, indicating higher DQE for the DR detectors and needle CR system than for the powder CR phosphor systems. The technical evaluation section of the study showed that the digital mammography systems were well set up and exhibiting typical performance for the detector technology employed in the respective systems. PMID:21701051

  7. Optimization of radiation dose and image quality in mammography: a clinical evaluation of rhodium versus molybdenum.

    PubMed

    Monticciolo, D L; Sprawls, P; Kruse, B D; Peterson, J E

    1996-04-01

    Mammography was done with a conventional molybdenum anode/filter combination and either a molybdenum/rhodium (Mo/Rh) or rhodium/rhodium (Rh/Rh) combination to compare the contrast characteristics and penetrating ability of the different techniques. Each pair was rated as being equal, one slightly better than the other, or one much better than the other. The "preferred" image was also selected. The entrance surface exposure and the mean glandular dose were determined for each exposure. Mo/Mo was compared with Mo/Rh in 26 patients and to Rh/Rh in 23. Mo/Mo was judged to produce better or equal contrast and penetration in all cases except one using the Mo/Rh combination. Mo/Rh was the preferred image only once and Rh/Rh not at all. In comparison to the conventional Mo/Mo technique, the average dose was reduced by approximately 25% with Mo/Rh and by 50% with Rh/Rh. PMID:8614878

  8. Authenticity and integrity of digital mammography images.

    PubMed

    Zhou, X Q; Huang, H K; Lou, S L

    2001-08-01

    Data security becomes more and more important in telemammography which uses a public high-speed wide area network connecting the examination site with the mammography expert center. Generally, security is characterized in terms of privacy, authenticity and integrity of digital data. Privacy is a network access issue and is not considered in this paper. We present a method, authenticity and integrity of digital mammography, here which can meet the requirements of authenticity and integrity for mammography image (IM) transmission. The authenticity and integrity for mammography (AIDM) consists of the following four modules. 1) Image preprocessing: To segment breast pixels from background and extract patient information from digital imaging and communication in medicine (DICOM) image header. 2) Image hashing: To compute an image hash value of the mammogram using the MD5 hash algorithm. 3) Data encryption: To produce a digital envelope containing the encrypted image hash value (digital signature) and corresponding patient information. 4) Data embedding: To embed the digital envelope into the image. This is done by replacing the least significant bit of a random pixel of the mammogram by one bit of the digital envelope bit stream and repeating for all bits in the bit stream. Experiments with digital IMs demonstrate the following. 1) In the expert center, only the user who knows the private key can open the digital envelope and read the patient information data and the digital signature of the mammogram transmitted from the examination site. 2) Data integrity can be verified by matching the image hash value decrypted from the digital signature with that computed from the transmitted image. 3) No visual quality degradation is detected in the embedded image compared with the original. Our preliminary results demonstrate that AIDM is an effective method for image authenticity and integrity in telemammography application. PMID:11513029

  9. Experimental evaluation of the image quality and dose in digital mammography: Influence of x-ray spectrum

    NASA Astrophysics Data System (ADS)

    Tomal, A.; Perez, A. M. M. M.; Silva, M. C.; Poletti, M. E.

    2015-11-01

    In this work, we studied experimentally the influence of x-ray spectrum on the contrast-to-noise ratio (CNR) and the average glandular dose (MDG) for two digital mammography systems: Senographe 2000D (GE Medical Systems) and Lorad Selenia (Hologic), with indirect and direct detector imaging technology, respectively. CNR and MGD were determined using PMMA phantoms simulating breasts with thicknesses of 4 cm and 6 cm. All available anode/filter combinations of the systems were evaluated for a wide range of tube voltages values. Results indicated that the Rh/Rh combination provides the highest image quality with the lower mean glandular dose for the Senographe 2000D system. For the Lorad Selenia system, the W/Ag combination at 30 kV showed the best performance, in terms of dose saving and image quality improvement in relation to all tube voltage range. The comparison between the optimal x-ray spectra and those selected by the AEC mode showed that this automatic selection mechanism could be readjusted to optimize the relationship between image quality and dose.

  10. Image quality evaluation of direct-conversion digital mammography system with new dual a-Se layer detector

    NASA Astrophysics Data System (ADS)

    Kuwabara, Takao; Iwasaki, Nobuyuki; Sendai, Tomonari; Furue, Ryosuke; Agano, Toshitaka

    2009-02-01

    To increase the detection performance of breast cancers in mammograms, we need to improve shape delineation of micro calcifications and tumors. We accomplished this by developing a direct-conversion mammography system with an optical reading method and a new dual a-Se layer detector. The system achieved both small pixel size (50 micrometer) and a high Detective Quantum Efficiency (DQE) realized by 100 % of fill factor and noise reduction. We evaluated image quality performance and determined the best exposure conditions. We measured DQE and Modulation Transfer Function(MTF) according to the IEC62220-1-2. High DQE was maintained at a low radiation dosage, indicating that the optical reading method accompanies low noises. Response of MTF was maintained at up to the Nyquist frequency of 10 cyc/mm, which corresponds to 50 micrometer pixel size. To determine the best exposure conditions, we measured Contrast to Noise Ratio (CNR) and visually evaluated images of a resected breast under conditions of MoMo, MoRh, and WRh. There were occasional disagreements between the exposure conditions for achieving the maximum CNR and those for the best image graded by the visual evaluation. This was probably because CNR measurement does not measure effects of scattered X-ray. The images verified the improvement in detection and delineation performance of micro calcifications and tumors.

  11. Dose sensitivity of three methods of image quality assessment in digital mammography

    NASA Astrophysics Data System (ADS)

    Hummel, Johann; Kaar, Marcus; Hoffmann, Rainer; Kaldarar, Heinrich; Semturs, Friedrich; Homolka, Peter; Figl, Michael

    2012-03-01

    Image quality assurance is one of the key issues in breast screening protocols. Although image quality can always be improved by increasing dose this mechanism is restricted by limiting values given by the standards. Therefore, it is crucial for system adjustment to describe the dependency of the image quality parameters on small changes in dose. This dose sensitivity was tested for three image quality evaluation methods. The European protocol requires the use of the CDMAM phantom which is a conventional contrast-detail phantom, while in North America the American College of Radiology (ACR) accreditation phantom is proposed. In contrast to these visual test methods the German PAS 1054 phantom uses digital image processing to derive image quality parameters like the noise-equivalent number of quanta (NEQ). We varied the dose within the range of clinical use. For the ACR phantom the examined parameter was the number of detected objects. With the CDMAM phantom we chose the diameters 0,10, 0.13, 0.20, 0.31 and 0.5 mm and recorded the threshold thicknesses. With respect to the PAS 1054 measurements we evaluated the NEQ at typical spatial frequencies to calculate the relative changes. NEQ versus dose increment shows a linear relationship and can be described by a linear function (R = .998). Every current-time product increment can be detected. With the ACR phantom the number of detected objects increases only in the lower dose range and reaches saturation at about 100mAs. The CDMAM can detect a 50% increase in dose confidently although the parameter increase is not monotonous. We conclude that an NEQ based method can be used as a simple and highly sensitive procedure for weekly quality assurance.

  12. Initial Image Quality and Clinical Experience with New CR Digital Mammography System: A Phantom and Clinical Study

    SciTech Connect

    Gaona, Enrique; Enriquez, Jesus Gabriel Franco; Alfonso, Beatriz Y. Alvarez; Castellanos, Gustavo Casian

    2008-08-11

    The goal of the study was to evaluate the first CR digital mammography system ( registered Konica-Minolta) in Mexico in clinical routine for cancer detection in a screening population and to determine if high resolution CR digital imaging is equivalent to state-of-the-art screen-film imaging. The mammograms were evaluated by two observers with cytological or histological confirmation for BIRADS 3, 4 and 5. Contrast, exposure and artifacts of the images were evaluated. Different details like skin, retromamillary space and parenchymal structures were judged. The detectability of microcalcifications and lesions were compared and correlated to histology. The difference in sensitivity of CR Mammography (CRM) and Screen Film Mammography (SFM) was not statistically significant. However, CRM had a significantly lower recall rate, and the lesion detection was equal or superior to conventional images. There is no significant difference in the number of microcalcifications and highly suspicious calcifications were equally detected on both film-screen and digital images. Different anatomical regions were better detectable in digital than in conventional mammography.

  13. CR mammography: Design and implementation of a quality control program

    SciTech Connect

    Moreno-Ramirez, A.; Brandan, M. E.; Villasenor-Navarro, Y.; Galvan, H. A.; Ruiz-Trejo, C.

    2012-10-23

    Despite the recent acquisition of significant quantities of computed radiography CR equipment for mammography, Mexican regulations do not specify the performance requirements for digital systems such as those of CR type. The design of a quality control program QCP specific for CR mammography systems was thus considered relevant. International protocols were taken as reference to define tests, procedures and acceptance criteria. The designed QCP was applied in three CR mammography facilities. Important deficiencies in spatial resolution, noise, image receptor homogeneity, artifacts and breast thickness compensation were detected.

  14. Mammography

    MedlinePLUS

    ... MRI) X-rays and other radiographic tests Mammography Nuclear medicine scans Ultrasound Categories of some common imaging tests Questions about radiation risk from imaging tests Factors that determine which ...

  15. Adaptation of a clustered lumpy background model for task-based image quality assessment in x-ray phase-contrast mammography

    PubMed Central

    Zysk, Adam M.; Brankov, Jovan G.; Wernick, Miles N.; Anastasio, Mark A.

    2012-01-01

    Purpose: Since the introduction of clinical x-ray phase-contrast mammography (PCM), a technique that exploits refractive-index variations to create edge enhancement at tissue boundaries, a number of optimization studies employing physical image-quality metrics have been performed. Ideally, task-based assessment of PCM would have been conducted with human readers. These studies have been limited, however, in part due to the large parameter-space of PCM system configurations and the difficulty of employing expert readers for large-scale studies. It has been proposed that numerical observers can be used to approximate the statistical performance of human readers, thus enabling the study of task-based performance over a large parameter-space. Methods: Methods are presented for task-based image quality assessment of PCM images with a numerical observer, the most significant of which is an adapted lumpy background from the conventional mammography literature that accounts for the unique wavefield propagation physics of PCM image formation and will be used with a numerical observer to assess image quality. These methods are demonstrated by performing a PCM task-based image quality study using a numerical observer. This study employs a signal-known-exactly, background-known-statistically Bayesian ideal observer method to assess the detectability of a calcification object in PCM images when the anode spot size and calcification diameter are varied. Results: The first realistic model for the structured background in PCM images has been introduced. A numerical study demonstrating the use of this background model has compared PCM and conventional mammography detection of calcification objects. The study data confirm the strong PCM calcification detectability dependence on anode spot size. These data can be used to balance the trade-off between enhanced image quality and the potential for motion artifacts that comes with use of a reduced spot size and increased exposure time. Conclusions: A method has been presented for the incorporation of structured breast background data into task-based numerical observer assessment of PCM images. The method adapts conventional background simulation techniques to the wavefield propagation physics necessary for PCM imaging. This method is demonstrated with a simple detection task. PMID:22320800

  16. Mammography

    MedlinePLUS

    ... breast cancer compared to film mammography. Three-dimensional (3D) mammography is a type of digital mammography. Researchers do not yet know whether 3D mammography is more or less accurate than standard ...

  17. Morphological breast imaging: tomography and digital mammography with synchrotron radiation

    NASA Astrophysics Data System (ADS)

    Longo, Renata; Pani, Silvia; Arfelli, Fulvia; Dreossi, Diego; Olivo, Alessandro; Poropat, Paolo; Quaia, Emilio; Rigon, Luigi; Zanconati, Fabrizio; Dalla Palma, Ludovico; Castelli, Edoardo

    2003-01-01

    A synchrotron radiation-based X-ray source offers a powerful tool for mammography due to the energy spectrum properties and the peculiar laminar beam geometry. Significant improvements in image quality have been achieved by the SYRMEP (SYnchrotron Radiation for MEdical Physics) collaboration, which has designed and built a beamline devoted to medical physics at the SR facility Elettra in Trieste (Italy). The detection system developed for digital mammography consists of a silicon pixel detector with 200×300 ?m 2 pixel size and high conversion efficiency. The detector is equipped with a low noise read-out electronics working in single photon counting mode. Mammographic phantoms and in vitro full breast samples have been investigated: the digital images show higher contrast resolution and lower absorbed dose than the images of the same samples obtained at the clinical mammographic unit.The SYRMEP collaboration is carrying out a breast tomography feasibility study to evaluate the image quality and the delivered dose. The SYRMEP beam is an ideal tool for tomography due to the laminar and monochromatic beam with negligible divergence. The experimental set-up and the acquisition protocol have been studied and the tomographic images of full breast samples acquired in the energy range 20-28 keV indicate that good quality images can be obtained with delivered doses comparable to conventional mammography.

  18. SU-E-I-88: Mammography Imaging: Does Positioning Matter?

    SciTech Connect

    Zhang, J; Szabunio, M

    2014-06-01

    Purpose: In mammography, compression is imperative for quality images and glandular radiation exposure dose. The thickness of the compressed breast directly determines mammography acquisition parameters. The compressed thickness varies due to variation in technologist practice, even for the same patient imaged at different time. This study is to investigate potential effect of the variation in breast positioning on radiation dose and image quality. Methods: Radiation dose at different thicknesses was measured with a BR-12 breast phantom for both conventional craniocaudal view and tomosynthesis in a Hologic Tomosynthesis mammography system. The CIRS stereotactic needle biopsy training phantom embedded dense masses and microcalcification in various sizes were imaged for image quality evaluation. Radiologists evaluated images. Clinical mammograms from the same patient but acquired at different time were retrospectively retrieved to evaluate potential effects of variation in positioning. Results: Acquisition parameters (kVp and mAs) increase with the increased phantom thickness. Radiation exposure increases following an exponential trend. The stereotactic phantom images showed loss of spatial and contrast resolution with inappropriate positioning. The compressed pressure may not be a good indicator for appropriate positioning. The inclusion of different amount of pectoralis muscle may lead to the same compressed pressure but different compressed thickness. The initial retrospective study of 3 patients showed that there were potential large variations in positioning the same patient at different examination time, resulting in large variations in patient radiation dose and image quality. Conclusion: Variations in patient positioning potentially influence patient radiation dose and image quality. The technologist has the critical responsibility to position patient to provide quality images in spite of different breast and body types. To reduce intra and inter practice variations in positioning patient, a training program among each breast imaging center may be a need.

  19. 76 FR 60848 - National Mammography Quality Assurance Advisory Committee; Notice of Meeting

    Federal Register 2010, 2011, 2012, 2013, 2014

    2011-09-30

    ... HUMAN SERVICES Food and Drug Administration National Mammography Quality Assurance Advisory Committee... be open to the public. Name of Committee: National Mammography Quality Assurance Advisory Committee...) Proposed changes to the Mammography Quality Standard Act (MQSA) policies and inspection procedures;...

  20. A task-based quality control metric for digital mammography

    NASA Astrophysics Data System (ADS)

    Maki Bloomquist, A. K.; Mainprize, J. G.; Mawdsley, G. E.; Yaffe, M. J.

    2014-11-01

    A reader study was conducted to tune the parameters of an observer model used to predict the detectability index (dʹ ) of test objects as a task-based quality control (QC) metric for digital mammography. A simple test phantom was imaged to measure the model parameters, namely, noise power spectrum, modulation transfer function and test-object contrast. These are then used in a non-prewhitening observer model, incorporating an eye-filter and internal noise, to predict dʹ. The model was tuned by measuring dʹ of discs in a four-alternative forced choice reader study. For each disc diameter, dʹ was used to estimate the threshold thicknesses for detectability. Data were obtained for six types of digital mammography systems using varying detector technologies and x-ray spectra. A strong correlation was found between measured and modeled values of dʹ, with Pearson correlation coefficient of 0.96. Repeated measurements from separate images of the test phantom show an average coefficient of variation in dʹ for different systems between 0.07 and 0.10. Standard deviations in the threshold thickness ranged between 0.001 and 0.017 mm. The model is robust and the results are relatively system independent, suggesting that observer model dʹ shows promise as a cross platform QC metric for digital mammography.

  1. Comparison of Acquisition Parameters and Breast Dose in Digital Mammography and Screen-Film Mammography in the American College of Radiology Imaging Network Digital Mammographic Imaging Screening Trial

    PubMed Central

    Hendrick, R. Edward; Pisano, Etta D.; Averbukh, Alice; Moran, Catherine; Berns, Eric A.; Yaffe, Martin J.; Herman, Benjamin; Acharyya, Suddhasatta; Gatsonis, Constantine

    2010-01-01

    OBJECTIVE The purpose of our study was to compare the technical performance of full-field digital mammography (FFDM) and screen-film mammography. MATERIALS AND METHODS The American College of Radiology Imaging Network Digital Mammographic Imaging Screening Trial enrolled 49,528 women to compare FFDM and screen-film mammography for screening. For quality assurance purposes, technical parameters including breast compression force, compressed breast thickness, mean glandular dose, and the number of additional views needed for complete breast coverage were recorded and analyzed for both FFDM and screen-film mammography on approximately 10% of study subjects at each site. RESULTS Technical data were compiled on 5,102 study subjects at 33 sites. Clean data were obtained for 4,366 (88%) of those cases. Mean compression force was 10.7 dN for screen-film mammography and 10.1 dN for FFDM (5.5% difference, p < 0.001). Mean compressed breast thickness was 5.3 cm for screen-film mammography and 5.4 cm for FFDM (1.7% difference, p < 0.001). Mean glandular dose per view averaged 2.37 mGy for screen-film mammography and 1.86 mGy for FFDM, 22% lower for digital than screen-film mammography, with sizeable variations among digital manufacturers. Twelve percent of screen-film mammography cases required more than the normal four views, whereas 21% of FFDM cases required more than the four normal views to cover all breast tissue. When extra views were included, mean glandular dose per subject was 4.15 mGy for FFDM and 4.98 mGy for screen-film mammography, 17% lower for FFDM than screen-film mammography. CONCLUSION Our results show that differences between screen-film mammography and FFDM in compression force and indicated compressed breast thickness were small. On average, FFDM had 22% lower mean glandular dose than screen-film mammography per acquired view, with sizeable variations in average FFDM doses by manufacturer. PMID:20093597

  2. Quality control for digital mammography: Part II recommendations from the ACRIN DMIST trial

    SciTech Connect

    Yaffe, Martin J.; Bloomquist, Aili K.; Mawdsley, Gordon E.

    2006-03-15

    The Digital Mammography Imaging Screening Trial (DMIST), conducted under the auspices of the American College of Radiology Imaging Network (ACRIN), is a clinical trial designed to compare the accuracy of digital versus screen-film mammography in a screening population [E. Pisano et al., ACRIN 6652--Digital vs. Screen-Film Mammography, ACRIN (2001)]. Part I of this work described the Quality Control program developed to ensure consistency and optimal operation of the digital equipment. For many of the tests, there were no failures during the 24 months imaging was performed in DMIST. When systems failed, they generally did so suddenly rather than through gradual deterioration of performance. In this part, the utility and effectiveness of those tests are considered. This suggests that after verification of proper operation, routine extensive testing would be of minimal value. A recommended set of tests is presented including additional and improved tests, which we believe meet the intent and spirit of the Mammography Quality Standards Act regulations to ensure that full-field digital mammography systems are functioning correctly, and consistently producing mammograms of excellent image quality.

  3. Full Field Digital Mammography (FFDM) versus CMOS Technology versus Tomosynthesis (DBT) – Which System Increases the Quality of Intraoperative Imaging?

    PubMed Central

    Schulz-Wendtland, R.; Dilbat, G.; Bani, M.; Fasching, P. A.; Lux, M. P.; Wenkel, E.; Schwab, S.; Loehberg, C. R.; Jud, S. M.; Rauh, C.; Bayer, C. M.; Beckmann, M. W.; Uder, M.; Meier-Meitinger, M.

    2012-01-01

    Aim: The aim of this prospective clinical study was to assess whether it would be possible to reduce the rate of re-excisions and improve the quality using CMOS technology or digital breast tomosynthesis (DBT) compared to a conventional FFDM system. Material and Methods: An invasive breast cancer (BI-RADS 5) was diagnosed in 200 patients in the period from 5/2011 to 1/2012. After histological verification, a breast-conserving therapy was performed with intraoperative imaging. Three different imaging systems were used: 1) Inspiration™ (Siemens, Erlangen, Germany), amorphous selenium, tungsten source, focus 0.1?mm, resolution 85?µm pixel pitch, 8?l/mm as the standard; 2) BioVision™ (Bioptics, Tucson, USA), flat panel photodiode array, tungsten source, focus 0.05, resolution 50?µm pixel pitch, 12?l/mm; 3) Tomosynthesis (Siemens, Erlangen, Germany), amorphous selenium, tungsten source, focus 0.1?mm, resolution 85?µm pixel pitch, 8?l/mm, range: 50°, 25 projections, scan time >?20?s, geometry: uniform scanning, reconstruction: filtered back projection. The 600 radiograms were prospectively shown to 3 radiologists. Results: Out of a total of 200 patients with histologically confirmed breast cancer (BI-RADS 6) 156 patients required no further operative therapy (re-excision) after breast-conserving therapy. A retrospective analysis (n?=?44) showed an increase in sensitivity with tomosynthesis compared to the BioVision™ (CMOS technology) and the Inspiration™ at a magnification of 1.0?:?1.0 of 8?% (p?

  4. Digital mammography, cancer screening: Factors important for image compression

    NASA Technical Reports Server (NTRS)

    Clarke, Laurence P.; Blaine, G. James; Doi, Kunio; Yaffe, Martin J.; Shtern, Faina; Brown, G. Stephen; Winfield, Daniel L.; Kallergi, Maria

    1993-01-01

    The use of digital mammography for breast cancer screening poses several novel problems such as development of digital sensors, computer assisted diagnosis (CAD) methods for image noise suppression, enhancement, and pattern recognition, compression algorithms for image storage, transmission, and remote diagnosis. X-ray digital mammography using novel direct digital detection schemes or film digitizers results in large data sets and, therefore, image compression methods will play a significant role in the image processing and analysis by CAD techniques. In view of the extensive compression required, the relative merit of 'virtually lossless' versus lossy methods should be determined. A brief overview is presented here of the developments of digital sensors, CAD, and compression methods currently proposed and tested for mammography. The objective of the NCI/NASA Working Group on Digital Mammography is to stimulate the interest of the image processing and compression scientific community for this medical application and identify possible dual use technologies within the NASA centers.

  5. 75 FR 70011 - Guidance for Industry, Mammography Quality Standards Act Inspectors, and Food and Drug...

    Federal Register 2010, 2011, 2012, 2013, 2014

    2010-11-16

    ...The Food and Drug Administration (FDA) is announcing the availability of the guidance entitled ``The Mammography Quality Standards Act Final Regulations: Modifications and Additions to Policy Guidance Help System 13.'' This document is intended to assist mammography facilities and their personnel in meeting the requirements of the Mammography Quality Standards Act (MQSA)...

  6. Computational assessment of mammography accreditation phantom images and correlation with human observer analysis

    NASA Astrophysics Data System (ADS)

    Barufaldi, Bruno; Lau, Kristen C.; Schiabel, Homero; Maidment, D. A.

    2015-03-01

    Routine performance of basic test procedures and dose measurements are essential for assuring high quality of mammograms. International guidelines recommend that breast care providers ascertain that mammography systems produce a constant high quality image, using as low a radiation dose as is reasonably achievable. The main purpose of this research is to develop a framework to monitor radiation dose and image quality in a mixed breast screening and diagnostic imaging environment using an automated tracking system. This study presents a module of this framework, consisting of a computerized system to measure the image quality of the American College of Radiology mammography accreditation phantom. The methods developed combine correlation approaches, matched filters, and data mining techniques. These methods have been used to analyze radiological images of the accreditation phantom. The classification of structures of interest is based upon reports produced by four trained readers. As previously reported, human observers demonstrate great variation in their analysis due to the subjectivity of human visual inspection. The software tool was trained with three sets of 60 phantom images in order to generate decision trees using the software WEKA (Waikato Environment for Knowledge Analysis). When tested with 240 images during the classification step, the tool correctly classified 88%, 99%, and 98%, of fibers, speck groups and masses, respectively. The variation between the computer classification and human reading was comparable to the variation between human readers. This computerized system not only automates the quality control procedure in mammography, but also decreases the subjectivity in the expert evaluation of the phantom images.

  7. Automated analysis of phantom images for the evaluation of long-term reproducibility in digital mammography

    NASA Astrophysics Data System (ADS)

    Gennaro, G.; Ferro, F.; Contento, G.; Fornasin, F.; di Maggio, C.

    2007-03-01

    The performance of an automatic software package was evaluated with phantom images acquired by a full-field digital mammography unit. After the validation, the software was used, together with a Leeds TORMAS test object, to model the image acquisition process. Process modelling results were used to evaluate the sensitivity of the method in detecting changes of exposure parameters from routine image quality measurements in digital mammography, which is the ultimate purpose of long-term reproducibility tests. Image quality indices measured by the software included the mean pixel value and standard deviation of circular details and surrounding background, contrast-to-noise ratio and relative contrast; detail counts were also collected. The validation procedure demonstrated that the software localizes the phantom details correctly and the difference between automatic and manual measurements was within few grey levels. Quantitative analysis showed sufficient sensitivity to relate fluctuations in exposure parameters (kVp or mAs) to variations in image quality indices. In comparison, detail counts were found less sensitive in detecting image quality changes, even when limitations due to observer subjectivity were overcome by automatic analysis. In conclusion, long-term reproducibility tests provided by the Leeds TORMAS phantom with quantitative analysis of multiple IQ indices have been demonstrated to be effective in predicting causes of deviation from standard operating conditions and can be used to monitor stability in full-field digital mammography.

  8. Evaluation of clinical image processing algorithms used in digital mammography.

    PubMed

    Zanca, Federica; Jacobs, Jurgen; Van Ongeval, Chantal; Claus, Filip; Celis, Valerie; Geniets, Catherine; Provost, Veerle; Pauwels, Herman; Marchal, Guy; Bosmans, Hilde

    2009-03-01

    Screening is the only proven approach to reduce the mortality of breast cancer, but significant numbers of breast cancers remain undetected even when all quality assurance guidelines are implemented. With the increasing adoption of digital mammography systems, image processing may be a key factor in the imaging chain. Although to our knowledge statistically significant effects of manufacturer-recommended image processings have not been previously demonstrated, the subjective experience of our radiologists, that the apparent image quality can vary considerably between different algorithms, motivated this study. This article addresses the impact of five such algorithms on the detection of clusters of microcalcifications. A database of unprocessed (raw) images of 200 normal digital mammograms, acquired with the Siemens Novation DR, was collected retrospectively. Realistic simulated microcalcification clusters were inserted in half of the unprocessed images. All unprocessed images were subsequently processed with five manufacturer-recommended image processing algorithms (Agfa Musica 1, IMS Raffaello Mammo 1.2, Sectra Mamea AB Sigmoid, Siemens OPVIEW v2, and Siemens OPVIEW v1). Four breast imaging radiologists were asked to locate and score the clusters in each image on a five point rating scale. The free-response data were analyzed by the jackknife free-response receiver operating characteristic (JAFROC) method and, for comparison, also with the receiver operating characteristic (ROC) method. JAFROC analysis revealed highly significant differences between the image processings (F = 8.51, p < 0.0001), suggesting that image processing strongly impacts the detectability of clusters. Siemens OPVIEW2 and Siemens OPVIEW1 yielded the highest and lowest performances, respectively. ROC analysis of the data also revealed significant differences between the processing but at lower significance (F = 3.47, p = 0.0305) than JAFROC. Both statistical analysis methods revealed that the same six pairs of modalities were significantly different, but the JAFROC confidence intervals were about 32% smaller than ROC confidence intervals. This study shows that image processing has a significant impact on the detection of microcalcifications in digital mammograms. Objective measurements, such as described here, should be used by the manufacturers to select the optimal image processing algorithm. PMID:19378737

  9. An SVM Based Approach for the Analysis Of Mammography Images

    NASA Astrophysics Data System (ADS)

    Gan, X.; Kapsokalivas, L.; Skaliotis, A.; Steinhöfel, K.; Tangaro, S.

    2007-09-01

    Mammography is among the most popular imaging techniques used in the diagnosis of breast cancer. Nevertheless distinguishing between healthy and ill images is hard even for an experienced radiologist, because a single image usually includes several regions of interest (ROIs). The hardness of this classification problem along with the substantial amount of data, gathered from patients' medical history, motivates the use of a machine learning approach as part of a CAD (Computer Aided Detection) tool, aiming to assist radiologists in the characterization of mammography images. Specifically, our approach involves: i) the ROI extraction, ii) the Feature Vector extraction, iii) the Support Vector Machine (SVM) classification of ROIs and iv) the characterization of the whole image. We evaluate the performance of our approach in terms of the SVM's training and testing error and in terms of ROI specificity—sensitivity. The results show a relation between the number of features used and the SVM's performance.

  10. An SVM Based Approach for the Analysis Of Mammography Images

    SciTech Connect

    Gan, X.; Kapsokalivas, L.; Skaliotis, A.; Steinhoefel, K.; Tangaro, S.

    2007-09-06

    Mammography is among the most popular imaging techniques used in the diagnosis of breast cancer. Nevertheless distinguishing between healthy and ill images is hard even for an experienced radiologist, because a single image usually includes several regions of interest (ROIs). The hardness of this classification problem along with the substantial amount of data, gathered from patients' medical history, motivates the use of a machine learning approach as part of a CAD (Computer Aided Detection) tool, aiming to assist radiologists in the characterization of mammography images. Specifically, our approach involves: i) the ROI extraction, ii) the Feature Vector extraction, iii) the Support Vector Machine (SVM) classification of ROIs and iv) the characterization of the whole image. We evaluate the performance of our approach in terms of the SVM's training and testing error and in terms of ROI specificity - sensitivity. The results show a relation between the number of features used and the SVM's performance.

  11. Comparing the Performance of Image Enhancement Methods to Detect Microcalcification Clusters in Digital Mammography

    PubMed Central

    Moradmand, Hajar; Setayeshi, Saeed; Karimian, Ali Reza; Sirous, Mehri; Akbari, Mohammad Esmaeil

    2012-01-01

    Background Mammography is the primary imaging technique for detection and diagnosis of breast cancer; however, the contrast of a mammogram image is often poor, especially for dense and glandular tissues. In these cases the radiologist may miss some diagnostically important microcalcifications. In order to improve diagnosis of cancer correctly, image enhancement technology is often used to enhance the image and help radiologists. Methods This paper presents a comparative study in digital mammography image enhancement based on four different algorithms: wavelet-based enhancement (Asymmetric Daubechies of order 8), Contrast-Limited Adaptive Histogram Equalization (CLAHE), morphological operators and unsharp masking. These algorithms have been tested on 114 clinical digital mammography images. The comparison for all the proposed image enhancement techniques was carried out to find out the best technique in enhancement of the mammogram images to detect microcalcifications. Results For evaluation of performance of image enhancement algorithms, the Contrast Improvement Index (CII) and profile intensity surface area distribution curve quality assessment have been used after any enhancement. The results of this study have shown that the average of CII is about 2.61 for wavelet and for CLAHE, unsharp masking and morphology operation are about 2.047, 1.63 and 1.315 respectively. Conclusion Experimental results strongly suggest that the wavelet transformation can be more effective and improve significantly overall detection of the Computer-Aided Diagnosis (CAD) system especially for dense breast. Compare to other studies, our method achieved a higher CII. PMID:25628822

  12. Ultra-Fast Image Reconstruction of Tomosynthesis Mammography Using GPU.

    PubMed

    Arefan, D; Talebpour, A; Ahmadinejhad, N; Kamali Asl, A

    2015-06-01

    Digital Breast Tomosynthesis (DBT) is a technology that creates three dimensional (3D) images of breast tissue. Tomosynthesis mammography detects lesions that are not detectable with other imaging systems. If image reconstruction time is in the order of seconds, we can use Tomosynthesis systems to perform Tomosynthesis-guided Interventional procedures. This research has been designed to study ultra-fast image reconstruction technique for Tomosynthesis Mammography systems using Graphics Processing Unit (GPU). At first, projections of Tomosynthesis mammography have been simulated. In order to produce Tomosynthesis projections, it has been designed a 3D breast phantom from empirical data. It is based on MRI data in its natural form. Then, projections have been created from 3D breast phantom. The image reconstruction algorithm based on FBP was programmed with C++ language in two methods using central processing unit (CPU) card and the Graphics Processing Unit (GPU). It calculated the time of image reconstruction in two kinds of programming (using CPU and GPU). PMID:26171373

  13. Ultra-Fast Image Reconstruction of Tomosynthesis Mammography Using GPU

    PubMed Central

    Arefan, D.; Talebpour, A.; Ahmadinejhad, N.; Kamali Asl, A.

    2015-01-01

    Digital Breast Tomosynthesis (DBT) is a technology that creates three dimensional (3D) images of breast tissue. Tomosynthesis mammography detects lesions that are not detectable with other imaging systems. If image reconstruction time is in the order of seconds, we can use Tomosynthesis systems to perform Tomosynthesis-guided Interventional procedures. This research has been designed to study ultra-fast image reconstruction technique for Tomosynthesis Mammography systems using Graphics Processing Unit (GPU). At first, projections of Tomosynthesis mammography have been simulated. In order to produce Tomosynthesis projections, it has been designed a 3D breast phantom from empirical data. It is based on MRI data in its natural form. Then, projections have been created from 3D breast phantom. The image reconstruction algorithm based on FBP was programmed with C++ language in two methods using central processing unit (CPU) card and the Graphics Processing Unit (GPU). It calculated the time of image reconstruction in two kinds of programming (using CPU and GPU). PMID:26171373

  14. Getting started with protocol for quality assurance of digital mammography in the clinical centre of Montenegro.

    PubMed

    Ivanovic, S; Bosmans, H; Mijovic, S

    2015-07-01

    The purpose of this work is (i) to work out a test procedure for quality assurance (QA) in digital mammography with newly released test equipment, including the MagicMax mam multimeter (IBA, Germany) and the anthropomorphic tissue equivalent phantom Mammo AT (IBA, Germany), and (ii) to determine whether a first digital computer radiography (CR) system in Montenegro meets the current European standards. Tested parameters were tube output (µGy mAs(-1)) and output rate (mGy s(-1)), reproducibility and accuracy of tube voltage, half value layer, reproducibility and accuracy of the AEC system, exposure control steps, image receptor's response function, image quality and printer stability test. The evaluated dosimetric quantity is the average glandular dose (AGD) as evaluated from PMMA slabs simulating breast tissue. The main findings are that QA can be organised in Montenegro. (1) All measured parameters are within the range described in European protocols except the tube voltage which deviated more than ± 1 kV. The automatic determination of the HVL was satisfactorily. AGD ranged from 0.66 to 7.02 mGy for PMMA thicknesses from 20 to 70 mm, and is in accordance with literature data. (2) The image quality score as obtained with the anthropomorphic tissue equivalent phantom Mammo AT for the CR system was similar to findings on the authors' conventional screen-film mammography. (3) In clinical practice the mammograms are printed. The CR reader produces images with a pixel size of 43.75 µm, which is compatible with the laser printer (39 µm laser spot spacing). The image processing algorithm embedded in the reader successfully processes mammograms with desirable image brightness and contrast in the printed image. The authors conclude that this first digital mammography system seems a good candidate for breast cancer screening applications. PMID:25862535

  15. Validation of MTF measurement for digital mammography quality control

    SciTech Connect

    Carton, Ann-Katherine; Vandenbroucke, Dirk; Struye, Luc; Maidment, Andrew D.A.; Kao, Y.-H.; Albert, Michael; Bosmans, Hilde; Marchal, Guy

    2005-06-15

    The modulation transfer function (MTF) describes the spatial resolution properties of imaging systems. In this work, the accuracy of our implementation of the edge method for calculating the presampled MTF was examined. Synthetic edge images with known MTF were used as gold standards for determining the robustness of the edge method. These images simulated realistic data from clinical digital mammography systems, and contained intrinsic system factors that could affect the MTF accuracy, such as noise, scatter, and flat-field nonuniformities. Our algorithm is not influenced by detector dose variations for MTF accuracy up to 1/2 the sampling frequency. We investigated several methods for noise reduction, including truncating the supersampled line spread function (LSF), windowing the LSF, applying a local exponential fit to the LSF, and applying a monotonic constraint to the supersampled edge spread function. Only the monotonic constraint did not introduce a systematic error; the other methods could result in MTF underestimation. Overall, our edge method consistently computed MTFs which were in good agreement with the true MTF. The edge method was then applied to images from a commercial storage-phosphor based digital mammography system. The calculated MTF was affected by the size (sides of 2.5, 5, or 10 cm) and the composition (lead or tungsten) of the edge device. However, the effects on the MTF were observed only with regard to the low frequency drop (LFD). Scatter nonuniformity was dependent on edge size, and could lead to slight underestimation of LFD. Nevertheless, this negative effect could be minimized by using an edge of 5 cm or larger. An edge composed of lead is susceptible to L-fluorescence, which causes overestimation of the LFD. The results of this work are intended to underline the need for clear guidelines if the MTF is to be given a more crucial role in acceptance tests and routine assessment of digital mammography systems: the MTF algorithm and edge object test tool need to be publicly validated.

  16. Image segmentation and 3D visualization for MRI mammography

    NASA Astrophysics Data System (ADS)

    Li, Lihua; Chu, Yong; Salem, Angela F.; Clark, Robert A.

    2002-05-01

    MRI mammography has a number of advantages, including the tomographic, and therefore three-dimensional (3-D) nature, of the images. It allows the application of MRI mammography to breasts with dense tissue, post operative scarring, and silicon implants. However, due to the vast quantity of images and subtlety of difference in MR sequence, there is a need for reliable computer diagnosis to reduce the radiologist's workload. The purpose of this work was to develop automatic breast/tissue segmentation and visualization algorithms to aid physicians in detecting and observing abnormalities in breast. Two segmentation algorithms were developed: one for breast segmentation, the other for glandular tissue segmentation. In breast segmentation, the MRI image is first segmented using an adaptive growing clustering method. Two tracing algorithms were then developed to refine the breast air and chest wall boundaries of breast. The glandular tissue segmentation was performed using an adaptive thresholding method, in which the threshold value was spatially adaptive using a sliding window. The 3D visualization of the segmented 2D slices of MRI mammography was implemented under IDL environment. The breast and glandular tissue rendering, slicing and animation were displayed.

  17. Digital mammography, sestamibi breast scintigraphy, and positron emission tomography breast imaging.

    PubMed

    Pisano, E D; Parham, C A

    2000-07-01

    Digital mammography allows for the separate optimization of image acquisition and display. Through this technology, and the application of image processing and computer aided diagnosis, breast cancer detection and breast lesion diagnosis might be improved. Besides the obvious data storage, retrieval, and transmission advantages that digital mammography will allow, additional advances such as tomosynthesis, dual energy mammography and digital subtraction mammography are in development. The possible future utility of Sestamibi breast scintigraphy and breast imaging with positron emission tomography is also discussed. PMID:10943283

  18. Compositional breast imaging using a dual-energy mammography protocol

    SciTech Connect

    Laidevant, Aurelie D.; Malkov, Serghei; Flowers, Chris I.; Kerlikowske, Karla; Shepherd, John A.

    2010-01-15

    Purpose: Mammography has a low sensitivity in dense breasts due to low contrast between malignant and normal tissue confounded by the predominant water density of the breast. Water is found in both adipose and fibroglandular tissue and constitutes most of the mass of a breast. However, significant protein mass is mainly found in the fibroglandular tissue where most cancers originate. If the protein compartment in a mammogram could be imaged without the influence of water, the sensitivity and specificity of the mammogram may be improved. This article describes a novel approach to dual-energy mammography, full-field digital compositional mammography (FFDCM), which can independently image the three compositional components of breast tissue: water, lipid, and protein. Methods: Dual-energy attenuation and breast shape measures are used together to solve for the three compositional thicknesses. Dual-energy measurements were performed on breast-mimicking phantoms using a full-field digital mammography unit. The phantoms were made of materials shown to have similar x-ray attenuation properties of the compositional compartments. They were made of two main stacks of thicknesses around 2 and 4 cm. Twenty-six thickness and composition combinations were used to derive the compositional calibration using a least-squares fitting approach. Results: Very high accuracy was achieved with a simple cubic fitting function with root mean square errors of 0.023, 0.011, and 0.012 cm for the water, lipid, and protein thicknesses, respectively. The repeatability (percent coefficient of variation) of these measures was tested using sequential images and was found to be 0.5%, 0.5%, and 3.3% for water, lipid, and protein, respectively. However, swapping the location of the two stacks of the phantom on the imaging plate introduced further errors showing the need for more complete system uniformity corrections. Finally, a preliminary breast image is presented of each of the compositional compartments separately. Conclusions: FFDCM has been derived and exhibited good compositional thickness accuracy on phantoms. Preliminary breast images demonstrated the feasibility of creating individual compositional diagnostic images in a clinical environment.

  19. Method for position emission mammography image reconstruction

    DOEpatents

    Smith, Mark Frederick

    2004-10-12

    An image reconstruction method comprising accepting coincidence datat from either a data file or in real time from a pair of detector heads, culling event data that is outside a desired energy range, optionally saving the desired data for each detector position or for each pair of detector pixels on the two detector heads, and then reconstructing the image either by backprojection image reconstruction or by iterative image reconstruction. In the backprojection image reconstruction mode, rays are traced between centers of lines of response (LOR's), counts are then either allocated by nearest pixel interpolation or allocated by an overlap method and then corrected for geometric effects and attenuation and the data file updated. If the iterative image reconstruction option is selected, one implementation is to compute a grid Siddon retracing, and to perform maximum likelihood expectation maiximization (MLEM) computed by either: a) tracing parallel rays between subpixels on opposite detector heads; or b) tracing rays between randomized endpoint locations on opposite detector heads.

  20. [Examinations of rare-earth film-screen-systems for image detectors in mammography (author's transl)].

    PubMed

    Weberling, R

    1979-03-01

    Intensifying screens based on Gd2 O2 S : Tb were analysed for x-ray exposure reduction and resolution using in mammorgraphy. The contrast of these screens coupled with a single emulsion film was tested and the density dominante was determined. A resolution grid on this density exposured shows the detail resolution and judge the quality profile of these new film-screen-combinations. To all factors was compared a non-screen industrial film for long time processing. The application of Gadolinium intensifying screens in mammography allowed an exposure reduction of 75% at the same image quality like a nonscreen industrial film. PMID:432483

  1. Evaluation of software for reading images of the CDMAM test object to assess digital mammography systems

    NASA Astrophysics Data System (ADS)

    Young, Kenneth C.; Alsager, Abdulaziz; Oduko, Jennifer M.; Bosmans, Hilde; Verbrugge, Beatrijs; Geertse, Tanya; van Engen, Ruben

    2008-03-01

    European Guidelines for quality control in digital mammography specify minimum and achievable standards of image quality in terms of threshold contrast, based on readings of images of the CDMAM test object by human observers. However this is time-consuming and has large inter- and intra-observer error. To overcome these problems a software program (CDCOM) is available to automatically read CDMAM images. After some further analysis the automated measurements can be used to predict the threshold contrast for a typical observer. The results of threshold contrast determination by human observers at three different centres were compared against automated readings. These data provide a means of predicting average human performance using the automated reading software. The coefficient of variation in automatically determined threshold gold thickness was about 4% for detail sizes from 0.2 to 1.0mm when 8 images were analysed. The coefficient of variation was about 10% at a detail size of 0.1mm. Using larger numbers of images improved reproducibility for all detail sizes. A change in phantom design could greatly improve reproducibility for the smallest detail sizes. Greater consistency of phantom construction would also be desirable as one of the four phantoms tested was significantly different from the other three. Despite some limitations automated reading of CDMAM images can provide a reproducible means of assessing digital mammography systems against European Guidelines.

  2. Advanced Breast Cancer as Indicator of Quality Mammography

    NASA Astrophysics Data System (ADS)

    Gaona, Enrique

    2003-09-01

    Breast cancer is the most frequently diagnosed cancer and is the second leading cause of cancer death among women in the Mexican Republic. Mammography is the more important screening tool for detecting early breast cancer. Screening mammography involves taking x-rays from two views from each breast, typically from above (cranial-caudal view, CC) and from an oblique or angled view (mediolateral-oblique, MLO). The purpose of this study was to carry out an exploratory survey of the issue of patients with advanced breast cancer who have had a screening mammography. A general result of the survey is that 22.5% of all patients (102) with advanced breast cancer that participated in the study had previous screening mammography. But we should consider that 10% of breast cancers are not detected by mammography. Only 70% of the family doctors prescribed a diagnostic mammography when the first symptoms were diagnosed.

  3. Comparative Study Of Image Enhancement Algorithms For Digital And Film Mammography

    SciTech Connect

    Delgado-Gonzalez, A.

    2008-08-11

    Here we discuss the application of edge enhancement algorithms on images obtained with a Mammography System which has a Selenium Detector and on the other hand, on images obtained from digitized film mammography. Comparative analysis of such images includes the study of technical aspects of image acquisition, storage, compression and display. A protocol for a local database has been created as a result of this study.

  4. Investigating the visual inspection subjectivity on the contrast-detail evaluation in digital mammography images

    NASA Astrophysics Data System (ADS)

    Sousa, Maria A. Z.; Medeiros, Regina B.; Schiabel, Homero

    2014-03-01

    A major difficulty in the interpretation of mammographic images is the low contrast and, in the case of early detection of breast cancer, the reduced size of the features of malignancy on findings such as microcalcifications. Furthermore, image assessment is subject to significant reliance of the capacity of observation of the expert that will perform it, compromising the final diagnosis accuracy. Thinking about this aspect, this study evaluated the subjectivity of visual inspection to assess the contrast-detail in mammographic images. For this, we compared the human readings of images generated with the CDMAM phantom performed by four observers, enabling to determining a threshold of contrast visibility in each diameter disks present in the phantom. These thresholds were compared graphically and by statistical measures allowing us to build a strategy for use of contrast and detail (dimensions) as parameters of quality in mammography.

  5. [Mammography screening. Concept, quality assurance and interdisciplinary cooperation].

    PubMed

    Heywang-Köbrunner, S H; Nährig, J; Hacker, A

    2008-11-01

    In 2005/2006 the German National Mammography Screening Program was initiated and has now become established. The objective is to reduce breast cancer mortality and the early diagnosis and therapy of small cancers. The program follows the European guidelines and is controlled by over 30 parameters of quality. All trained members of the team document each step of the screening chain electronically. Histological assessment (HA) is recommended in up to 2% of examinations, 90% of HAs are performed by core needle biopsy (CNB) or by stereotactic vacuum-assisted biopsy (VABB). Open diagnostic biopsies are performed in <10% of all HAs and therapy is successful in some of the B3 lesions. Mammograms are interpreted by two independent readers. Recommendations of the regular interdisciplinary conferences, preoperative and postoperative, follow the European guidelines. About 45% of all breast cancers detected by screening are in-situ or less than 10 mm in size. The 17% alterations diagnosed by needle biopsy are B3 or B4 lesions and impose high demands on the pathologists and the interdisciplinary team. Due to the many early and discrete lesions counterchecking of representative biopsies is crucial. Problems may be caused by sampling error or partial volume effects. Interdisciplinary conferences and knowledge of the limitations of each discipline and method are needed to optimize diagnostic and therapeutic decisions. PMID:18807041

  6. Stereotactic mammography imaging combined with 3D US imaging for image guided breast biopsy

    SciTech Connect

    Surry, K. J. M.; Mills, G. R.; Bevan, K.; Downey, D. B.; Fenster, A.

    2007-11-15

    Stereotactic X-ray mammography (SM) and ultrasound (US) guidance are both commonly used for breast biopsy. While SM provides three-dimensional (3D) targeting information and US provides real-time guidance, both have limitations. SM is a long and uncomfortable procedure and the US guided procedure is inherently two dimensional (2D), requiring a skilled physician for both safety and accuracy. The authors developed a 3D US-guided biopsy system to be integrated with, and to supplement SM imaging. Their goal is to be able to biopsy a larger percentage of suspicious masses using US, by clarifying ambiguous structures with SM imaging. Features from SM and US guided biopsy were combined, including breast stabilization, a confined needle trajectory, and dual modality imaging. The 3D US guided biopsy system uses a 7.5 MHz breast probe and is mounted on an upright SM machine for preprocedural imaging. Intraprocedural targeting and guidance was achieved with real-time 2D and near real-time 3D US imaging. Postbiopsy 3D US imaging allowed for confirmation that the needle was penetrating the target. The authors evaluated 3D US-guided biopsy accuracy of their system using test phantoms. To use mammographic imaging information, they registered the SM and 3D US coordinate systems. The 3D positions of targets identified in the SM images were determined with a target localization error (TLE) of 0.49 mm. The z component (x-ray tube to image) of the TLE dominated with a TLE{sub z} of 0.47 mm. The SM system was then registered to 3D US, with a fiducial registration error (FRE) and target registration error (TRE) of 0.82 and 0.92 mm, respectively. Analysis of the FRE and TRE components showed that these errors were dominated by inaccuracies in the z component with a FRE{sub z} of 0.76 mm and a TRE{sub z} of 0.85 mm. A stereotactic mammography and 3D US guided breast biopsy system should include breast compression for stability and safety and dual modality imaging for target localization. The system will provide preprocedural x-ray mammography information in the form of SM imaging along with real-time US imaging for needle guidance to a target. 3D US imaging will also be available for targeting, guidance, and biopsy verification immediately postbiopsy.

  7. An Assessment of the Quality of Mammography Care at Facilities Treating Medically Vulnerable Populations

    PubMed Central

    Goldman, L. Elizabeth; Haneuse, Sebastien J.-P. A.; Miglioretti, Diana L.; Kerlikowske, Karla; Buist, Diana S. M.; Yankaskas, Bonnie; Smith-Bindman, Rebecca

    2009-01-01

    Background Women in medically vulnerable populations, including racial and ethnic minorities, the socioeconomically disadvantaged, and residents of rural areas, experience higher breast cancer mortality than do others. Whether mammography facilities that treat vulnerable women demonstrate lower quality of care than other facilities is unknown. Objectives To assess the quality of mammography women receive at facilities characterized as serving a high proportion of medically vulnerable populations. Research Design We prospectively collected self-reported breast cancer risk factor information, mammography interpretations, and cancer outcomes on 1,579,929 screening mammography examinations from 750,857 women, aged 40–80 years, attending any of 151 facilities in the Breast Cancer Surveillance Consortium, between 1998 and 2004. To classify facilities as serving medically vulnerable populations, we used 4 criteria: educational attainment, racial/ethnic minority, household income, and rural/urban residence. Results After adjustment for patient-level factors known to effect mammography accuracy, facilities serving vulnerable populations had significantly higher mammography specificity than did other facilities: ie, those serving women who were minorities [odds ratio (OR): 1.32; 95% confidence interval (CI): 1.01–1.73], living in rural areas (1.45; 1.15–1.73), and with lower household income (1.33; 1.05–1.68). We observed no statistically significant differences between facilities in mammography sensitivity. Conclusions Facilities serving high proportions of vulnerable populations provide screening mammography with equal or better quality (as reflected in higher specificity with no corresponding decrease in sensitivity) than other facilities. Further research is needed to understand the mechanisms underlying these findings. PMID:18580389

  8. Positron emission mammography: high-resolution biochemical breast imaging.

    PubMed

    Weinberg, Irving N; Beylin, David; Zavarzin, Valera; Yarnall, Steve; Stepanov, Pavel Y; Anashkin, Edward; Narayanan, Deepa; Dolinsky, Sergei; Lauckner, Kathrin; Adler, Lee P

    2005-02-01

    Positron emission mammography (PEM) provides images of biochemical activity in the breast with spatial resolution matching individual ducts (1.5 mm full-width at half-maximum). This spatial resolution, supported by count efficiency that results in high signal-to-noise ratio, allows confident visualization of intraductal as well as invasive breast cancers. Clinical trials with a full-breast PEM device have shown high clinical accuracy in characterizing lesions identified as suspicious on the basis of conventional imaging or physical examination (sensitivity 93%, specificity 83%, area under the ROC curve of 0.93), with high sensitivity preserved (91%) for intraductal cancers. Increased sensitivity did not come at a cost of reduced specificity. Considering that intraductal cancer represents more than 30% of reported cancers, and is the form of cancer with the highest probability of achieving surgical cure, it is likely that the use of PEM will complement anatomic imaging modalities in the areas of surgical planning, high-risk monitoring, and minimally invasive therapy. The quantitative nature of PET promises to assist researchers interested studying the response of putative cancer precursors (e.g., atypical ductal hyperplasia) to candidate prevention agents. PMID:15649088

  9. On imaging with or without grid in digital mammography

    NASA Astrophysics Data System (ADS)

    Chen, Han; Danielsson, Mats; Cederström, Björn

    2014-03-01

    The grids used in digital mammography to reduce scattered radiation from the breast are not perfect and lead to partial absorption of primary radiation at the same time as not all of the scattered radiation is absorbed. It has therefore lately been suggested to remove the grids and correct for effects of scattered radiation by post- processing the images. In this paper, we investigated the dose reduction that might be achieved if the gird were to be removed. Dose reduction is determined as a function of PMMA thickness by comparing the contrast-to-noise ratios (CNRs) of images acquired with and without grid at a constant exposure. We used a theoretical model validated with Monte Carlo simulations and phantom studies. To evaluate the CNR, we applied aluminum filters of two different sizes, 4x8 cm2 and 1x1 cm2. When the large Al filter was used, the resulting CNR value for the grid-less images was overestimated as a result of a difference in amount of scattered radiation in the background region and of the region covered by the filter, a difference that could be eliminated by selecting a region of interest close to the edge of the filter. The optimal CNR when the PMMA thickness was above about 4 cm was obtained with a grid, whereas removing the grid leaded to a dose saving in thinner PMMAs. The results suggest not removing grids in breast cancer screening.

  10. Preoperative digital mammography imaging in conservative mastectomy and immediate reconstruction

    PubMed Central

    Angrigiani, Claudio; Hammond, Dennis; Nava, Maurizio; Gonzalez, Eduardo; Rostagno, Roman; Gercovich, Gustavo

    2016-01-01

    Background Digital mammography clearly distinguishes gland tissue density from the overlying non-glandular breast tissue coverage, which corresponds to the existing tissue between the skin and the Cooper’s ligaments surrounding the gland (i.e., dermis and subcutaneous fat). Preoperative digital imaging can determine the thickness of this breast tissue coverage, thus facilitating planning of the most adequate surgical techniques and reconstructive procedures for each case. Methods This study aimed to describe the results of a retrospective study of 352 digital mammograms in 176 patients with different breast volumes who underwent preoperative conservative mastectomies. The breast tissue coverage thickness and its relationship with the breast volume were evaluated. Results The breast tissue coverage thickness ranged from 0.233 to 4.423 cm, with a mean value of 1.952 cm. A comparison of tissue coverage and breast volume revealed a non-direct relationship between these factors. Conclusions Preoperative planning should not depend only on breast volume. Flap evaluations based on preoperative imaging measurements might be helpful when planning a conservative mastectomy. Accordingly, we propose a breast tissue coverage classification (BTCC). PMID:26855903

  11. Digital Mammography

    Cancer.gov

    Conventional mammography uses X-rays to look for tumors or suspicious areas in the breasts. Digital mammography also uses X-rays, but the data is collected on computer instead of on a piece of film. This means that the image can be computer-enhanced,

  12. Dosimetry and kVp standardization for quality assurance of mammography

    NASA Astrophysics Data System (ADS)

    Chu, Chien-Hau; Yuan, Ming-Chen; Huang, Wen-Sheng; Hsieh, Bor-Tsung

    2014-11-01

    Breast cancer mortality rates were significantly reduced in Taiwan after achieving early-stage monitoring with mammography screening. This study establishes an appropriate and traceable calibration infrastructure, which offers calibration services for mammography X-ray quality assurance instrumentation, which is performed clinically on a regular basis. The entrance air kerma, HVL, and kVp of mammography equipment with five different target/filter combinations can be taken as adequate indicators for the level of average glandular dose (AGD). The primary dose standard in mammography uses a free-air ionization chamber to estimate the rate of air kerma. Several correction factors were determined by Monte Carlo simulations and experiments. A secondary kVp standard in mammography is in accordance with the IEC 61676 recommendations. The calibration system of kVp meter uses a high-voltage divider, which is traceable to ITRI primary standard in Taiwan. Dose and kVp verifications were conducted by mammography instruments, which were previously calibrated by NIST and PTB. The evaluation results indicate that the capabilities of this irradiation system met the ISO 4037-1 requirements. The expanded uncertainties (k=2) were 1.03% and 1.6% when the mammography X-ray air kerma rate and kVp meter calibration factors were evaluated using ISO GUM. Experimental verification and a comparison with NIST using transfer ionization chambers yielded differences in calibration factors. Comparison with the PTB using kVp meter indicated a less than 1% difference. The results showed that dose and kVp standards were in reasonable agreement with standard uncertainty. The low uncertainties associated with the obtained results in this work show that the standardization employed can be accurately used for calibration of instrument in mammography in Taiwan.

  13. Development and validation of a modelling framework for simulating 2D-mammography and breast tomosynthesis images.

    PubMed

    Elangovan, Premkumar; Warren, Lucy M; Mackenzie, Alistair; Rashidnasab, Alaleh; Diaz, Oliver; Dance, David R; Young, Kenneth C; Bosmans, Hilde; Strudley, Celia J; Wells, Kevin

    2014-08-01

    Planar 2D x-ray mammography is generally accepted as the preferred screening technique used for breast cancer detection. Recently, digital breast tomosynthesis (DBT) has been introduced to overcome some of the inherent limitations of conventional planar imaging, and future technological enhancements are expected to result in the introduction of further innovative modalities. However, it is crucial to understand the impact of any new imaging technology or methodology on cancer detection rates and patient recall. Any such assessment conventionally requires large scale clinical trials demanding significant investment in time and resources. The concept of virtual clinical trials and virtual performance assessment may offer a viable alternative to this approach. However, virtual approaches require a collection of specialized modelling tools which can be used to emulate the image acquisition process and simulate images of a quality indistinguishable from their real clinical counterparts. In this paper, we present two image simulation chains constructed using modelling tools that can be used for the evaluation of 2D-mammography and DBT systems. We validate both approaches by comparing simulated images with real images acquired using the system being simulated. A comparison of the contrast-to-noise ratios and image blurring for real and simulated images of test objects shows good agreement ( < 9% error). This suggests that our simulation approach is a promising alternative to conventional physical performance assessment followed by large scale clinical trials. PMID:25029333

  14. Practice policy and quality initiatives: using lean principles to improve screening mammography workflow.

    PubMed

    Shah, Carla J; Sullivan, Julie R; Gonyo, Mary Beth; Wadhwa, Anubha; DuBois, Melissa S

    2013-01-01

    The "lean" approach is a quality improvement method that focuses on maximizing activities that are valued by the customer and eliminating waste that impedes efficiency in the workplace. The unique philosophy of the lean approach encourages all members of the team to be directly involved in identifying areas of waste and generating solutions to eliminate them. When the breast imaging section at the authors' institution became part of a multispecialty breast care center, the result was escalating examination volumes, more complex cases, and overall increased demand on radiologists' time. After several unsuccessful attempts to improve the efficiency of the section, including evaluation by outside consultants, the decision was made to embark on a comprehensive quality improvement program using the lean approach. A team of radiologists, technologists, file room personnel, information technology (IT) representatives, and administrators from the breast imaging section met twice a month to learn about lean principles and how to apply them to screening mammography workflows. Sources of inefficiency (waste) were identified, and potential solutions were generated. Multiple trials were performed to test these solutions. Throughout the process, all team members were engaged in identifying the problems, suggesting solutions, and implementing change. Most of the tested solutions were successful and resulted in decreased patient wait times, improved efficiency for the technologists and radiologists, faster report turnaround, and advances in IT. In addition, staff members were introduced to the lean philosophy and became actively involved in improving their workplace, resulting in a more cohesive section. PMID:23813321

  15. Attenuation correction with region growing method used in the positron emission mammography imaging system

    NASA Astrophysics Data System (ADS)

    Gu, Xiao-Yue; Li, Lin; Yin, Peng-Fei; Yun, Ming-Kai; Chai, Pei; Huang, Xian-Chao; Sun, Xiao-Li; Wei, Long

    2015-10-01

    The Positron Emission Mammography imaging system (PEMi) provides a novel nuclear diagnosis method dedicated for breast imaging. With a better resolution than whole body PET, PEMi can detect millimeter-sized breast tumors. To address the requirement of semi-quantitative analysis with a radiotracer concentration map of the breast, a new attenuation correction method based on a three-dimensional seeded region growing image segmentation (3DSRG-AC) method has been developed. The method gives a 3D connected region as the segmentation result instead of image slices. The continuity property of the segmentation result makes this new method free of activity variation of breast tissues. The threshold value chosen is the key process for the segmentation method. The first valley in the grey level histogram of the reconstruction image is set as the lower threshold, which works well in clinical application. Results show that attenuation correction for PEMi improves the image quality and the quantitative accuracy of radioactivity distribution determination. Attenuation correction also improves the probability of detecting small and early breast tumors. Supported by Knowledge Innovation Project of The Chinese Academy of Sciences (KJCX2-EW-N06)

  16. What's new in mammography.

    PubMed

    Simonetti, G; Cossu, E; Montanaro, M; Caschili, C; Giuliani, V

    1998-05-01

    Early diagnosis of breast cancer plays the leading role in reducing mortality rates and improving the patients' prognosis: mammography is the most sensitive technique currently available for the detection of nonpalpable lesions and therefore the method of choice. However, mammography has some limitations and the technique must be improved with technological devices without affecting image quality. This could be the target to increase diagnostic accuracy. Mammography sensitivity and specificity are now improved with the digital computer assisted technique, teleradiology, digital tomosynthesis or digital angiography--used to study microvascularization--3D imaging or synchrotron light, and laser mammography. Such other technological devices as Mammospot reduce breast thickness and provide better breast compression. Digital mammography can be carried out with film or direct digitization. The advantages of the digital technique are a shorter examination time, less storage space, electronic image recording, with image 'adjustments' made by the radiologist, and especially computerized analysis. The computer aided diagnosis can be defined as the diagnosis made by the radiologist who considers the results of computerized analysis as a 'second opinion'. In this way incidental mistakes made by radiologists, can be corrected by the computer analysis. Computers are a basic element also in teleradiology, which needs immediate and simultaneous admittance to the patient's history and permits radiology optimization in rural areas too. As for tomosynthesis, it permits to study a single slice of the breast without glandular tissue overlapping, which is useful in dense breasts where the diagnosis can be made with a lower X-ray dose. Moreover, this method fits the current mammographic systems easily. 3D imaging is still a work in progress. Synchrotron mammography is used only on surgery specimens, where it exhibits high resolution and contrast, depicting structures and details missed by conventional mammography. Breast DSA allows the study of vessels < 0.20 mm in diameter and of fine microvascular details; it can also demonstrate neoangiogenesis. Laser mammography permits bilateral examinations of the breast in 10-15 mins and is currently used also for breast cancer therapy, although only in animal trials. To conclude, after reviewing new techniques and evaluating the real cost/benefit ratio for each of them, conventional mammography remains the most sensitive tool for breast cancer diagnosis. PMID:9652528

  17. Task-based strategy for optimized contrast enhanced breast imaging: analysis of six imaging techniques for mammography and tomosynthesis

    NASA Astrophysics Data System (ADS)

    Ikejimba, Lynda; Kiarashi, Nooshin; Lin, Yuan; Chen, Baiyu; Ghate, Sujata V.; Zerhouni, Moustafa; Samei, Ehsan; Lo, Joseph Y.

    2012-03-01

    Digital breast tomosynthesis (DBT) is a novel x-ray imaging technique that provides 3D structural information of the breast. In contrast to 2D mammography, DBT minimizes tissue overlap potentially improving cancer detection and reducing number of unnecessary recalls. The addition of a contrast agent to DBT and mammography for lesion enhancement has the benefit of providing functional information of a lesion, as lesion contrast uptake and washout patterns may help differentiate between benign and malignant tumors. This study used a task-based method to determine the optimal imaging approach by analyzing six imaging paradigms in terms of their ability to resolve iodine at a given dose: contrast enhanced mammography and tomosynthesis, temporal subtraction mammography and tomosynthesis, and dual energy subtraction mammography and tomosynthesis. Imaging performance was characterized using a detectability index d', derived from the system task transfer function (TTF), an imaging task, iodine contrast, and the noise power spectrum (NPS). The task modeled a 5 mm lesion containing iodine concentrations between 2.1 mg/cc and 8.6 mg/cc. TTF was obtained using an edge phantom, and the NPS was measured over several exposure levels, energies, and target-filter combinations. Using a structured CIRS phantom, d' was generated as a function of dose and iodine concentration. In general, higher dose gave higher d', but for the lowest iodine concentration and lowest dose, dual energy subtraction tomosynthesis and temporal subtraction tomosynthesis demonstrated the highest performance.

  18. A Reconstruction Algorithm for Breast Cancer Imaging With Electrical Impedance Tomography in Mammography Geometry

    PubMed Central

    Kao, Tzu-Jen; Isaacson, David; Saulnier, Gary J.; Newell, Jonathan C.

    2009-01-01

    The conductivity and permittivity of breast tumors are known to differ significantly from those of normal breast tissues, and electrical impedance tomography (EIT) is being studied as a modality for breast cancer imaging to exploit these differences. At present, X-ray mammography is the primary standard imaging modality used for breast cancer screening in clinical practice, so it is desirable to study EIT in the geometry of mammography. This paper presents a forward model of a simplified mammography geometry and a reconstruction algorithm for breast tumor imaging using EIT techniques. The mammography geometry is modeled as a rectangular box with electrode arrays on the top and bottom planes. A forward model for the electrical impedance imaging problem is derived for a homogeneous conductivity distribution and is validated by experiment using a phantom tank. A reconstruction algorithm for breast tumor imaging based on a linearization approach and the proposed forward model is presented. It is found that the proposed reconstruction algorithm performs well in the phantom experiment, and that the locations of a 5-mm-cube metal target and a 6-mm-cube agar target could be recovered at a target depth of 15 mm using a 32 electrode system. PMID:17405377

  19. Task-based strategy for optimized contrast enhanced breast imaging: Analysis of six imaging techniques for mammography and tomosynthesis

    SciTech Connect

    Ikejimba, Lynda C.; Kiarashi, Nooshin; Ghate, Sujata V.; Samei, Ehsan; Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University Medical Center, Durham, North Carolina 27705; Department of Electrical and Computer Engineering, Duke University, Durham, North Carolina 27705; Department of Physics, Duke University, Durham, North Carolina 27705; Department of Biomedical Engineering, Duke University, Durham, North Carolina 27705 ; Lo, Joseph Y.; Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University Medical Center, Durham, North Carolina 27705; Department of Electrical and Computer Engineering, Duke University, Durham, North Carolina 27705; Department of Biomedical Engineering, Duke University, Durham, North Carolina 27705

    2014-06-15

    Purpose: The use of contrast agents in breast imaging has the capability of enhancing nodule detectability and providing physiological information. Accordingly, there has been a growing trend toward using iodine as a contrast medium in digital mammography (DM) and digital breast tomosynthesis (DBT). Widespread use raises concerns about the best way to use iodine in DM and DBT, and thus a comparison is necessary to evaluate typical iodine-enhanced imaging methods. This study used a task-based observer model to determine the optimal imaging approach by analyzing six imaging paradigms in terms of their ability to resolve iodine at a given dose: unsubtracted mammography and tomosynthesis, temporal subtraction mammography and tomosynthesis, and dual energy subtraction mammography and tomosynthesis. Methods: Imaging performance was characterized using a detectability index d{sup ′}, derived from the system task transfer function (TTF), an imaging task, iodine signal difference, and the noise power spectrum (NPS). The task modeled a 10 mm diameter lesion containing iodine concentrations between 2.1 mg/cc and 8.6 mg/cc. TTF was obtained using an edge phantom, and the NPS was measured over several exposure levels, energies, and target-filter combinations. Using a structured CIRS phantom, d{sup ′} was generated as a function of dose and iodine concentration. Results: For all iodine concentrations and dose, temporal subtraction techniques for mammography and tomosynthesis yielded the highest d{sup ′}, while dual energy techniques for both modalities demonstrated the next best performance. Unsubtracted imaging resulted in the lowest d{sup ′} values for both modalities, with unsubtracted mammography performing the worst out of all six paradigms. Conclusions: At any dose, temporal subtraction imaging provides the greatest detectability, with temporally subtracted DBT performing the highest. The authors attribute the successful performance to excellent cancellation of inplane structures and improved signal difference in the lesion.

  20. Task-based strategy for optimized contrast enhanced breast imaging: Analysis of six imaging techniques for mammography and tomosynthesis

    PubMed Central

    Ikejimba, Lynda C.; Kiarashi, Nooshin; Ghate, Sujata V.; Samei, Ehsan; Lo, Joseph Y.

    2014-01-01

    Purpose: The use of contrast agents in breast imaging has the capability of enhancing nodule detectability and providing physiological information. Accordingly, there has been a growing trend toward using iodine as a contrast medium in digital mammography (DM) and digital breast tomosynthesis (DBT). Widespread use raises concerns about the best way to use iodine in DM and DBT, and thus a comparison is necessary to evaluate typical iodine-enhanced imaging methods. This study used a task-based observer model to determine the optimal imaging approach by analyzing six imaging paradigms in terms of their ability to resolve iodine at a given dose: unsubtracted mammography and tomosynthesis, temporal subtraction mammography and tomosynthesis, and dual energy subtraction mammography and tomosynthesis. Methods: Imaging performance was characterized using a detectability index d?, derived from the system task transfer function (TTF), an imaging task, iodine signal difference, and the noise power spectrum (NPS). The task modeled a 10 mm diameter lesion containing iodine concentrations between 2.1 mg/cc and 8.6 mg/cc. TTF was obtained using an edge phantom, and the NPS was measured over several exposure levels, energies, and target-?lter combinations. Using a structured CIRS phantom, d? was generated as a function of dose and iodine concentration. Results: For all iodine concentrations and dose, temporal subtraction techniques for mammography and tomosynthesis yielded the highest d?, while dual energy techniques for both modalities demonstrated the next best performance. Unsubtracted imaging resulted in the lowest d? values for both modalities, with unsubtracted mammography performing the worst out of all six paradigms. Conclusions: At any dose, temporal subtraction imaging provides the greatest detectability, with temporally subtracted DBT performing the highest. The authors attribute the successful performance to excellent cancellation of inplane structures and improved signal difference in the lesion. PMID:24877819

  1. REVIEW: The development of mammography

    NASA Astrophysics Data System (ADS)

    Law, J.

    2006-07-01

    This review traces the development of mammography physics over the last 50 years, concentrating mainly on technological changes and their inter-relations. It has been written for physicists with no specific mammography experience but a general interest in radiology, as much as for those with recent involvement in mammography. Topics covered include industrial film, xerography, intensifying screens, x-ray tube developments, image quality test objects, patient dose and performance checks. Some of these developments were necessary before population screening of healthy women could be considered, while others have resulted from increased opportunities for equipment manufactures which screening programmes created. The standpoint of this review is that of a physicist with long experience in a UK centre where mammography was performed on dedicated equipment well over 40 years ago and where screening has been performed continuously for 30 years.

  2. Experimental and Monte Carlo-simulated spectra of standard mammography-quality beams

    PubMed Central

    David, M G; Pires, E J; Bernal, M A; Peixoto, J G; Dealmeida, C E

    2012-01-01

    ABSTRACT A spectrometric study of standard mammography-quality beams by using experimental and Monte Carlo simulation methods was carried out in this work. The qualities of these beams are described according to the International Electrotechical Commission 61267 standard and the Technical Report Series 457 International Atomic Energy Agency report. Specifically, the non-attenuated RQR-M beam series was studied. Methods A Si-PIN diode-based spectrometer and the PENELOPE Monte Carlo code (v. 2008F1) were used for experiments and simulations, respectively. In addition, an ionization chamber was used to determine the half-value layers (HVLs) of each beam quality. The measurements were done in the mammography dosimeter calibration setup of our laboratory, and the Monte Carlo simulations reproduced such conditions. Results The relative differences between the HVLs calculated from experimental and simulated spectra were lower than 2.4% for all the beam qualities studied. These differences are 1.2% and 3.1% when comparing the HVLs calculated from the experimental and simulated spectra to those determined by using the ionization chamber, respectively. A semi-empirical relation was found to obtain the nominal tube potential from the effective tube potential. Conclusion According to our results, the mammography beams used in this work have energy spectra similar to clinical beams. PMID:22010026

  3. Three-dimensional Breast Imaging with Full Field Digital Mammography Tomosynthesis

    NASA Astrophysics Data System (ADS)

    Eberhard, Jeffrey W.

    2003-03-01

    Although conventional film-screen mammography is the clinical modality of choice for early detection of breast cancer, many cancers are missed because they are masked by radiographically dense fibroglandular breast tissue which may be overlying or surrounding the tumor. The superposition of 3D breast anatomy in a standard 2D x-ray projection is perhaps the most significant problem in mammography today. GE Global Research has developed a new 3D full field digital mammography tomosynthesis prototype system that directly addresses the superimposed tissue problem by enabling volumetric imaging of the breast. High performance digital detectors with low electronic noise and fast read-out times, new reconstruction algorithms customized for tomosynthesis acquisitions, and application of volume rendering methods to enable rapid, effective review of 3D data are among the key enabling technologies for tomosynthesis. Phantom studies have demonstrated significantly enhanced performance of tomosynthesis compared to standard digital mammography exams. Over 200 patients have been imaged with a prototype system. Typical patient images will be shown.

  4. CADx Mammography

    NASA Astrophysics Data System (ADS)

    Costaridou, Lena

    Although a wide variety of Computer-Aided Diagnosis (CADx) schemes have been proposed across breast imaging modalities, and especially in mammography, research is still ongoing to meet the high performance CADx requirements. In this chapter, methodological contributions to CADx in mammography and adjunct breast imaging modalities are reviewed, as they play a major role in early detection, diagnosis and clinical management of breast cancer. At first, basic terms and definitions are provided. Then, emphasis is given to lesion content derivation, both anatomical and functional, considering only quantitative image features of micro-calcification clusters and masses across modalities. Additionally, two CADx application examples are provided. The first example investigates the effect of segmentation accuracy on micro-calcification cluster morphology derivation in X-ray mammography. The second one demonstrates the efficiency of texture analysis in quantification of enhancement kinetics, related to vascular heterogeneity, for mass classification in dynamic contrast-enhanced magnetic resonance imaging.

  5. Digital mammography.

    PubMed

    Pisano, Etta D; Yaffe, Martin J

    2005-02-01

    In digital mammography, the processes of image acquisition, display, and storage are separated, which allows optimization of each. Radiation transmitted through the breast is absorbed by an electronic detector, the response of which is faithful over a wide range of intensities. Once this information is recorded, it can be displayed by using computer image-processing techniques to allow arbitrary settings of image brightness and contrast, without the need for further exposure to the patient. In this article, the current state of the art in technology for digital mammography and data from clinical trials that support the use of the technology will be reviewed. In addition, several potentially useful applications that are being developed with digital mammography will be described. PMID:15670993

  6. Synchrotron Radiation Mammography: Clinical Experimentation

    SciTech Connect

    Arfelli, Fulvia; Dreossi, Diego; Longo, Renata; Rokvic, Tatjana; Castelli, Edoardo; Abrami, Alessandro; Chenda, Valentina; Menk, Ralf-Hendrik; Quai, Elisa; Tromba, Giuliana; Bregant, Paola; De Guarrini, Fabio; Cova, Maria A.; Tonutti, Maura; Zanconati, Fabrizio

    2007-01-19

    For several years a large variety of in-vitro medical imaging studies were carried out at the SYRMEP (Synchrotron Radiation for Medical Physics) beamline of the synchrotron radiation facility ELETTRA (Trieste, Italy) utilizing phase sensitive imaging techniques. In particular low dose Phase Contrast (PhC) in planar imaging mode and computed tomography were utilized for full field mammography. The results obtained on in-vitro samples at the SYRMEP beamline in PhC breast imaging were so encouraging that a clinical program on a limited number of patients selected by radiologists was launched to validate the improvements of synchrotron radiation in mammography. PhC mammography with conventional screen-film systems is the first step within this project. A digital system is under development for future applications. During the last years the entire beamline has been deeply modified and a medical facility dedicated to in-vivo mammography was constructed. The facility for PhC synchrotron radiation mammography is now operative in patient mode. The system reveals a prominent increase in image quality with respect to conventional mammograms even at lower delivered dose.

  7. Digital Breast Imaging Workflow.

    PubMed

    Odle, Teresa G

    2016-01-01

    Most breast imaging programs have replaced film-screen mammography systems with digital mammography. Other breast imaging methods, such as digital breast tomosynthesis, complement the tools available for screening and diagnosis of breast cancer. This article examines digital mammography, other digital breast imaging modalities, and the effects of these technologies on practice, quality, efficiency, and the technologist's role in patient care. PMID:26721854

  8. Optical mammography instrument for broadband spectral imaging with depth discrimination

    NASA Astrophysics Data System (ADS)

    Krishnamurthy, Nishanth; Kainerstorfer, Jana M.; Anderson, Pamela G.; Sassaroli, Angelo; Fantini, Sergio

    2015-03-01

    We are developing a new instrument for diffuse optical mammography in parallel plate geometry that operates over a broad spectral range of 600-1000 nm, features a scan time of 1-2 min, and allows for dynamic measurements at a selected region of interest. Furthermore, this new instrument is capable of depth discrimination of optical inhomogeneities embedded in the examined tissue by using multiple off-axis detection fibers. Using a solid silicone phantoms, mimicking breast tissue with 39 mm thickness, we demonstrate the capability of this instrument to recover the depth of blood-vessel-like structures to within ~2 mm. Additionally, we demonstrate the capability of this instrument to perform dynamic optical measurements with a temporal sampling rate as high as 20 Hz. We describe our plans to integrate this rich spectral, spatial, and temporal information into a single instrument for translation into clinical measurements on breast cancer patients.

  9. Measurements and simulations of scatter imaging as a simultaneous adjunct for screening mammography

    NASA Astrophysics Data System (ADS)

    Kern, Katie; Hassan, Laila; Peerzada, Lubna; Ur-Rehman, Mahboob; MacDonald, C. A.

    2015-03-01

    X-ray coherent scatter is dependent upon the molecular structure of the scattering material and hence allows differentiation between tissue types with potentially much higher contrast than conventional absorption-based radiography. Coherent-scatter computed tomography has been used to produce images based on the x-ray scattering properties of the tissue. However, the geometry for CT imaging requires a thin fan beam and multiple projections and is incommensurate with screening mammography. In this work we demonstrate progress in a developing a system using a wide slot beam and simple anti-scatter grid which is adequate to differentiate between scatter peaks to remove the fat background from the coherent scatter image. Adequate intensity in the coherent scatter image can be achieved at the dose commonly used for screening mammography to detect carcinoma surrogates as small as 2 mm in diameter. This technique would provide an inexpensive, low dose, simultaneous adjunct to conventional screening mammography to provide a localized map of tissue type that could be overlaid on the conventional transmission mammogram. Comparisons between phantom measurements and Monte Carlo simulations show good agreement, which allowed for detailed examination of the visibility of carcinoma under realistic conditions.

  10. The effect of image processing on the detection of cancers in digital mammography.

    PubMed

    Warren, Lucy M; Given-Wilson, Rosalind M; Wallis, Matthew G; Cooke, Julie; Halling-Brown, Mark D; Mackenzie, Alistair; Chakraborty, Dev P; Bosmans, Hilde; Dance, David R; Young, Kenneth C

    2014-08-01

    OBJECTIVE. The objective of our study was to investigate the effect of image processing on the detection of cancers in digital mammography images. MATERIALS AND METHODS. Two hundred seventy pairs of breast images (both breasts, one view) were collected from eight systems using Hologic amorphous selenium detectors: 80 image pairs showed breasts containing subtle malignant masses; 30 image pairs, biopsy-proven benign lesions; 80 image pairs, simulated calcification clusters; and 80 image pairs, no cancer (normal). The 270 image pairs were processed with three types of image processing: standard (full enhancement), low contrast (intermediate enhancement), and pseudo-film-screen (no enhancement). Seven experienced observers inspected the images, locating and rating regions they suspected to be cancer for likelihood of malignancy. The results were analyzed using a jackknife-alternative free-response receiver operating characteristic (JAFROC) analysis. RESULTS. The detection of calcification clusters was significantly affected by the type of image processing: The JAFROC figure of merit (FOM) decreased from 0.65 with standard image processing to 0.63 with low-contrast image processing (p = 0.04) and from 0.65 with standard image processing to 0.61 with film-screen image processing (p = 0.0005). The detection of noncalcification cancers was not significantly different among the image-processing types investigated (p > 0.40). CONCLUSION. These results suggest that image processing has a significant impact on the detection of calcification clusters in digital mammography. For the three image-processing versions and the system investigated, standard image processing was optimal for the detection of calcification clusters. The effect on cancer detection should be considered when selecting the type of image processing in the future. PMID:25055275

  11. Design of a novel phase contrast x-ray imaging system for mammography.

    PubMed

    Munro, Peter R T; Ignatyev, Konstantin; Speller, Robert D; Olivo, Alessandro

    2010-07-21

    It is hoped that x-ray phase contrast imaging (XPCi) will provide a generational improvement in the effectiveness of mammography. XPCi is sensitive to the refraction which x-rays undergo as a result of the variation in x-ray propagation speeds within an object. XPCi is, however, seldom used in clinical applications owing mainly to a lack of suitable systems. The radiation physics group at UCL has previously designed and built an XPCi system sensitive to phase gradients in one dimension for application in security inspection. We present here the design methodology and final design of a prototype XPCi system sensitive to phase gradients in two directions for use in mammography. The technique makes efficient use of the flux available from a laboratory x-ray source, thus making it suitable for clinical use. PMID:20601778

  12. Mammography accreditation program

    SciTech Connect

    Wilcox, P.

    1993-12-31

    In the mid-1980`s, the movement toward the use of dedicated mammography equipment provided significant improvement in breast cancer detection. However, several studies demonstrated that this change was not sufficient to ensure optimal image quality at a low radiation dose. In particular, the 1985 Nationwide Evaluation of X-ray Trends identified the wide variations in image quality and radiation dose, even from dedicated units. During this time period, the American Cancer Society (ACS) launched its Breast Cancer Awareness Screening Campaign. However, there were concerns about the ability of radiology to respond to the increased demand for optimal screening examinations that would result from the ACS program. To respond to these concerns, the ACS and the American College of Radiology (ACR) established a joint committee on mammography screening in 1986. After much discussion, it was decided to use the ACR Diagnostic Practice Accreditation Program as a model for the development of a mammography accreditation program. However, some constraints were required in order to make the program meet the needs of the ACS. This voluntary, peer review program had to be timely and cost effective. It was determined that the best way to address these needs would be to conduct the program by mail. Finally, by placing emphasis on the educational nature of the program, it would provide an even greater opportunity for improving mammographic quality. The result of this effort was that, almost six years ago, in May 1987, the pilot study for the ACR Mammography Accreditation Program (MAP) began, and in August of that year, the first applications were received. In November 1987, the first 3-year accreditation certificates were awarded.

  13. Complete internal audit of a mammography service in a reference institution for breast imaging*

    PubMed Central

    Badan, Gustavo Machado; Roveda Júnior, Décio; Ferreira, Carlos Alberto Pecci; de Noronha Junior, Ozeas Alves

    2014-01-01

    Objective Undertaking of a complete audit of the service of mammography, as recommended by BI-RADS®, in a private reference institution for breast cancer diagnosis in the city of São Paulo, SP, Brazil, and comparison of results with those recommended by the literature. Materials and Methods Retrospective, analytical and cross-sectional study including 8,000 patients submitted to mammography in the period between April 2010 and March 2011, whose results were subjected to an internal audit. The patients were followed-up until December 2012. Results The radiological classification of 7,249 screening mammograms, according to BI-RADS, was the following: category 0 (1.43%), 1 (7.82%), 2 (80.76%), 3 (8.35%), 4 (1.46%), 5 (0.15%) and 6 (0.03%). The breast cancer detection ratio was 4.8 cases per 1,000 mammograms. Ductal carcinoma in situ was found in 22.8% of cases. Positive predictive values for categories 3, 4 and 5 were 1.3%, 41.3% and 100%, respectively. In the present study, the sensitivity of the method was 97.1% and specificity, 97.4%. Conclusion The complete internal audit of a service of mammography is essential to evaluate the quality of such service, which reflects on an early breast cancer detection and reduction of mortality rates. PMID:25741052

  14. Computer-aided diagnostics of screening mammography using content-based image retrieval

    NASA Astrophysics Data System (ADS)

    Deserno, Thomas M.; Soiron, Michael; de Oliveira, Júlia E. E.; de A. Araújo, Arnaldo

    2012-03-01

    Breast cancer is one of the main causes of death among women in occidental countries. In the last years, screening mammography has been established worldwide for early detection of breast cancer, and computer-aided diagnostics (CAD) is being developed to assist physicians reading mammograms. A promising method for CAD is content-based image retrieval (CBIR). Recently, we have developed a classification scheme of suspicious tissue pattern based on the support vector machine (SVM). In this paper, we continue moving towards automatic CAD of screening mammography. The experiments are based on in total 10,509 radiographs that have been collected from different sources. From this, 3,375 images are provided with one and 430 radiographs with more than one chain code annotation of cancerous regions. In different experiments, this data is divided into 12 and 20 classes, distinguishing between four categories of tissue density, three categories of pathology and in the 20 class problem two categories of different types of lesions. Balancing the number of images in each class yields 233 and 45 images remaining in each of the 12 and 20 classes, respectively. Using a two-dimensional principal component analysis, features are extracted from small patches of 128 x 128 pixels and classified by means of a SVM. Overall, the accuracy of the raw classification was 61.6 % and 52.1 % for the 12 and the 20 class problem, respectively. The confusion matrices are assessed for detailed analysis. Furthermore, an implementation of a SVM-based CBIR system for CADx in screening mammography is presented. In conclusion, with a smarter patch extraction, the CBIR approach might reach precision rates that are helpful for the physicians. This, however, needs more comprehensive evaluation on clinical data.

  15. Optimal x-ray energy for digital mammography

    SciTech Connect

    Logan, C.M.; Hernandez, J.M.; Kinney, J.H.; Lewis, D.L.

    1992-11-01

    Screening mammography is a radiological procedure requiring the highest possible image quality at the lowest possible dose. It is widely recognized that digital image acquisition, computer assisted diagnosis, and scientific visualization can provide substantial improvement in mammography. For such systems, much of what is accepted as best practice with today`s film/screen/lightbox systems will become inappropriate. A complete system design is required. We have constructed a model of the breast imaging process. These results show that molybdenum-anode, molybdenum-filtered x-ray spectra are ill-suited for digital mammography. An x-ray spectrum rich in 22-to 25-keV photons is needed.

  16. Image quality and dose efficiency of high energy phase sensitive x-ray imaging: Phantom studies

    PubMed Central

    Wong, Molly Donovan; Wu, Xizeng; Liu, Hong

    2014-01-01

    The goal of this preliminary study was to perform an image quality comparison of high energy phase sensitive imaging with low energy conventional imaging at similar radiation doses. The comparison was performed with the following phantoms: American College of Radiology (ACR), contrast-detail (CD), acrylic edge and tissue-equivalent. Visual comparison of the phantom images indicated comparable or improved image quality for all phantoms. Quantitative comparisons were performed through ACR and CD observer studies, both of which indicated higher image quality in the high energy phase sensitive images. The results of this study demonstrate the ability of high energy phase sensitive imaging to overcome existing challenges with the clinical implementation of phase contrast imaging and improve the image quality for a similar radiation dose as compared to conventional imaging near typical mammography energies. In addition, the results illustrate the capability of phase sensitive imaging to sustain the image quality improvement at high x-ray energies and for – breast – simulating phantoms, both of which indicate the potential to benefit fields such as mammography. Future studies will continue to investigate the potential for dose reduction and image quality improvement provided by high energy phase sensitive contrast imaging. PMID:24865208

  17. Monte Carlo image simulation in support of improvements in mammography

    SciTech Connect

    Peplow, D.E.; Verghese, K.

    1997-12-01

    One in nine women in the United States is affected by breast cancer. Mammographic screenings currently miss {approximately}10% of palpable breast tumors in women over the age of 50; 25% are missed in younger women because of denser breast tissue. It is of vital interest to lower these percentages through contrast improvements in the images and to enable tumor detection at earlier stages, especially in younger women.

  18. Predicting diagnostic error in Radiology via eye-tracking and image analytics: Application in mammography

    SciTech Connect

    Voisin, Sophie; Pinto, Frank M; Morin-Ducote, Garnetta; Hudson, Kathy; Tourassi, Georgia

    2013-01-01

    Purpose: The primary aim of the present study was to test the feasibility of predicting diagnostic errors in mammography by merging radiologists gaze behavior and image characteristics. A secondary aim was to investigate group-based and personalized predictive models for radiologists of variable experience levels. Methods: The study was performed for the clinical task of assessing the likelihood of malignancy of mammographic masses. Eye-tracking data and diagnostic decisions for 40 cases were acquired from 4 Radiology residents and 2 breast imaging experts as part of an IRB-approved pilot study. Gaze behavior features were extracted from the eye-tracking data. Computer-generated and BIRADs images features were extracted from the images. Finally, machine learning algorithms were used to merge gaze and image features for predicting human error. Feature selection was thoroughly explored to determine the relative contribution of the various features. Group-based and personalized user modeling was also investigated. Results: Diagnostic error can be predicted reliably by merging gaze behavior characteristics from the radiologist and textural characteristics from the image under review. Leveraging data collected from multiple readers produced a reasonable group model (AUC=0.79). Personalized user modeling was far more accurate for the more experienced readers (average AUC of 0.837 0.029) than for the less experienced ones (average AUC of 0.667 0.099). The best performing group-based and personalized predictive models involved combinations of both gaze and image features. Conclusions: Diagnostic errors in mammography can be predicted reliably by leveraging the radiologists gaze behavior and image content.

  19. Predicting diagnostic error in radiology via eye-tracking and image analytics: Preliminary investigation in mammography

    SciTech Connect

    Voisin, Sophie; Tourassi, Georgia D.; Pinto, Frank; Morin-Ducote, Garnetta; Hudson, Kathleen B.

    2013-10-15

    Purpose: The primary aim of the present study was to test the feasibility of predicting diagnostic errors in mammography by merging radiologists’ gaze behavior and image characteristics. A secondary aim was to investigate group-based and personalized predictive models for radiologists of variable experience levels.Methods: The study was performed for the clinical task of assessing the likelihood of malignancy of mammographic masses. Eye-tracking data and diagnostic decisions for 40 cases were acquired from four Radiology residents and two breast imaging experts as part of an IRB-approved pilot study. Gaze behavior features were extracted from the eye-tracking data. Computer-generated and BIRADS images features were extracted from the images. Finally, machine learning algorithms were used to merge gaze and image features for predicting human error. Feature selection was thoroughly explored to determine the relative contribution of the various features. Group-based and personalized user modeling was also investigated.Results: Machine learning can be used to predict diagnostic error by merging gaze behavior characteristics from the radiologist and textural characteristics from the image under review. Leveraging data collected from multiple readers produced a reasonable group model [area under the ROC curve (AUC) = 0.792 ± 0.030]. Personalized user modeling was far more accurate for the more experienced readers (AUC = 0.837 ± 0.029) than for the less experienced ones (AUC = 0.667 ± 0.099). The best performing group-based and personalized predictive models involved combinations of both gaze and image features.Conclusions: Diagnostic errors in mammography can be predicted to a good extent by leveraging the radiologists’ gaze behavior and image content.

  20. Experience in reading digital images may decrease observer accuracy in mammography

    NASA Astrophysics Data System (ADS)

    Rawashdeh, Mohammad A.; Lewis, Sarah J.; Lee, Warwick; Mello-Thoms, Claudia; Reed, Warren M.; McEntee, Mark; Tapia, Kriscia; Brennan, Patrick C.

    2015-03-01

    Rationale and Objectives: To identify parameters linked to higher levels of performance in screening mammography. In particular we explored whether experience in reading digital cases enhances radiologists' performance. Methods: A total of 60 cases were presented to the readers, of which 20 contained cancers and 40 showed no abnormality. Each case comprised of four images and 129 breast readers participated in the study. Each reader was asked to identify and locate any malignancies using a 1-5 confidence scale. All images were displayed using 5MP monitors, supported by radiology workstations with full image manipulation capabilities. A jack-knife free-response receiver operating characteristic, figure of merit (JAFROC, FOM) methodology was employed to assess reader performance. Details were obtained from each reader regarding their experience, qualifications and breast reading activities. Spearman and Mann Whitney U techniques were used for statistical analysis. Results: Higher performance was positively related to numbers of years professionally qualified (r= 0.18; P<0.05), number of years reading breast images (r= 0.24; P<0.01), number of mammography images read per year (r= 0.28; P<0.001) and number of hours reading mammographic images per week (r= 0.19; P<0.04). Unexpectedly, higher performance was inversely linked to previous experience with digital images (r= - 0.17; p<0.05) and further analysis, demonstrated that this finding was due to changes in specificity. Conclusion: This study suggests suggestion that readers with experience in digital images reporting may exhibit a reduced ability to correctly identify normal appearances requires further investigation. Higher performance is linked to number of cases read per year.

  1. Comparison of radiation exposure and associated radiation-induced cancer risks from mammography and molecular imaging of the breast

    SciTech Connect

    O'Connor, Michael K.; Li Hua; Rhodes, Deborah J.; Hruska, Carrie B.; Clancy, Conor B.; Vetter, Richard J.

    2010-12-15

    Purpose: Recent studies have raised concerns about exposure to low-dose ionizing radiation from medical imaging procedures. Little has been published regarding the relative exposure and risks associated with breast imaging techniques such as breast specific gamma imaging (BSGI), molecular breast imaging (MBI), or positron emission mammography (PEM). The purpose of this article was to estimate and compare the risks of radiation-induced cancer from mammography and techniques such as PEM, BSGI, and MBI in a screening environment. Methods: The authors used a common scheme for all estimates of cancer incidence and mortality based on the excess absolute risk model from the BEIR VII report. The lifetime attributable risk model was used to estimate the lifetime risk of radiation-induced breast cancer incidence and mortality. All estimates of cancer incidence and mortality were based on a population of 100 000 females followed from birth to age 80 and adjusted for the fraction that survives to various ages between 0 and 80. Assuming annual screening from ages 40 to 80 and from ages 50 to 80, the cumulative cancer incidence and mortality attributed to digital mammography, screen-film mammography, MBI, BSGI, and PEM was calculated. The corresponding cancer incidence and mortality from natural background radiation was calculated as a useful reference. Assuming a 15%-32% reduction in mortality from screening, the benefit/risk ratio for the different imaging modalities was evaluated. Results: Using conventional doses of 925 MBq Tc-99m sestamibi for MBI and BSGI and 370 MBq F-18 FDG for PEM, the cumulative cancer incidence and mortality were found to be 15-30 times higher than digital mammography. The benefit/risk ratio for annual digital mammography was >50:1 for both the 40-80 and 50-80 screening groups, but dropped to 3:1 for the 40-49 age group. If the primary use of MBI, BSGI, and PEM is in women with dense breast tissue, then the administered doses need to be in the range 75-150 MBq for Tc-99m sestamibi and 35 MBq-70 MBq for F-18 FDG in order to obtain benefit/risk ratios comparable to those of mammography in these age groups. These dose ranges should be achievable with enhancements to current technology while maintaining a reasonable examination time. Conclusions: The results of the dose estimates in this study clearly indicate that if molecular imaging techniques are to be of value in screening for breast cancer, then the administered doses need to be substantially reduced to better match the effective doses of mammography.

  2. Investigation of physical image characteristics and phenomenon of edge enhancement by phase contrast using equipment typical for mammography

    SciTech Connect

    Yamazaki, Asumi; Ichikawa, Katsuhiro; Kodera, Yoshie

    2008-11-15

    A technique called phase contrast mammography (PCM) has only recently been applied in clinical examination. In this application, PCM images are acquired at a 1.75x magnification using an x-ray tube for clinical use, and then reduced to the real size of the object by image processing. The images showed enhanced object edges; reportedly, this enhancement occurred because of the refraction of x rays through a cylindrical object. The authors measured the physical image characteristics of PCM to compare the image characteristics of PCM with those of conventional mammography. More specifically, they measured the object-edge-response characteristics and the noise characteristics in the spatial frequency domain. The results revealed that the edge-response characteristics of PCM outperformed those of conventional mammography. In addition, the characteristics changed with the object-placement conditions and the object shapes. The noise characteristics of PCM were better than those of conventional mammography. Subsequently, to verify why object edges were enhanced in PCM images, the authors simulated image profiles that would be obtained if the x rays were refracted and totally reflected by using not only a cylindrical substance but also a planar substance as the object. So, they confirmed that the object edges in PCM images were enhanced because x rays were refracted irrespective of the object shapes. Further, they found that the edge enhancements depended on the object shapes and positions. It was also proposed that the larger magnification than 1.75 in the commercialized system might be more suitable for PCM. Finally, the authors investigated phase-contrast effects to breast tissues by the simulation and demonstrated that PCM would be helpful in the diagnoses of mammography.

  3. Co-registration of Positron Emission Mammography (PEM) images and x-ray mammograms

    SciTech Connect

    Bergman, A.M.; Robar, J.L.; Thompson, C.J.

    1996-12-31

    Accurate co-registration of metabolic and X-ray images of the breast is important when acquiring information about the location and diagnosis of suspicious lesions or tumors. X-ray mammograms reveal abnormal tissue densities, while metabolic images identify regions of abnormal metabolism. Conventional nuclear medicine and radiologic breast images must be acquired at different times with different patient positions making co-registration difficult. Positron Emission Mammography (PEM) simplifies the image registration process by acquiring both an X-ray density image and a metabolic image consecutively, without moving the patient between scans. We have integrated PEM detectors into a conventional mammographic unit and we have developed a co-registration tool. The tool is a thin sheet of plexiglass (130 x 95 x 3 mm) with a 49 x 39 mm{sup 2} window. The tool protrudes from the side of the upper PEM detector. The suspicious area of the breast is imaged through this tool during a magnified X-ray mammogram, producing a radio-lucent region on the resulting film image. During a positron emission scan, detectors acquire an image of the same region. The PEM detectors can be positioned anywhere along the width of the breast to provide a complete scan.

  4. Multiple-reader studies, digital mammography, computer-aided diagnosis, and the Holy Grail of imaging physics: I

    NASA Astrophysics Data System (ADS)

    Wagner, Robert F.; Beiden, Sergey V.; Campbell, Gregory

    2001-06-01

    There are multiple sources of variability in clinical studies of imaging systems. The variation of the reader `mindset' establishes the need for ROC analysis to control for that fundamental variable. The demonstration of the range of reader skills in mammography shows the need for a multivariate approach to ROC analysis. The multiple-reader, multiple-case (MRMC) ROC experimental paradigm addresses this need and several practical solutions to the problem of analysis of MRMC data have been developed. We review the application of these methods to an important clinical comparison of digital and conventional mammography.

  5. Predicting malignancy from mammography findings and image-guided core biopsies

    PubMed Central

    Ferreira, Pedro; Fonseca, Nuno A.; Dutra, Inês; Woods, Ryan; Burnside, Elizabeth

    2016-01-01

    The main goal of this work is to produce machine learning models that predict the outcome of a mammography from a reduced set of annotated mammography findings. In the study we used a dataset consisting of 348 consecutive breast masses that underwent image guided core biopsy performed between October 2005 and December 2007 on 328 female subjects. We applied various algorithms with parameter variation to learn from the data. The tasks were to predict mass density and to predict malignancy. The best classifier that predicts mass density is based on a support vector machine and has accuracy of 81.3%. The expert correctly annotated 70% of the mass densities. The best classifier that predicts malignancy is also based on a support vector machine and has accuracy of 85.6%, with a positive predictive value of 85%. One important contribution of this work is that our model can predict malignancy in the absence of the mass density attribute, since we can fill up this attribute using our mass density predictor. PMID:26333262

  6. Imaging performance of an amorphous selenium digital mammography detector in a breast tomosynthesis system

    SciTech Connect

    Zhao Bo; Zhao Wei

    2008-05-15

    In breast tomosynthesis a rapid sequence of N images is acquired when the x-ray tube sweeps through different angular views with respect to the breast. Since the total dose to the breast is kept the same as that in regular mammography, the exposure used for each image of tomosynthesis is 1/N. The low dose and high frame rate pose a tremendous challenge to the imaging performance of digital mammography detectors. The purpose of the present work is to investigate the detector performance in different operational modes designed for tomosynthesis acquisition, e.g., binning or full resolution readout, the range of view angles, and the number of views N. A prototype breast tomosynthesis system with a nominal angular range of {+-}25 deg. was used in our investigation. The system was equipped with an amorphous selenium (a-Se) full field digital mammography detector with pixel size of 85 {mu}m. The detector can be read out in full resolution or 2x1 binning (binning in the tube travel direction). The focal spot blur due to continuous tube travel was measured for different acquisition geometries, and it was found that pixel binning, instead of focal spot blur, dominates the detector modulation transfer function (MTF). The noise power spectrum (NPS) and detective quantum efficiency (DQE) of the detector were measured with the exposure range of 0.4-6 mR, which is relevant to the low dose used in tomosynthesis. It was found that DQE at 0.4 mR is only 20% less than that at highest exposure for both detector readout modes. The detector temporal performance was categorized as lag and ghosting, both of which were measured as a function of x-ray exposure. The first frame lags were 8% and 4%, respectively, for binning and full resolution mode. Ghosting is negligible and independent of the frame rate. The results showed that the detector performance is x-ray quantum noise limited at the low exposures used in each view of tomosynthesis, and the temporal performance at high frame rate (up to 2 frames per second) is adequate for tomosynthesis.

  7. Data acquisition and analysis of mammography images at the NSLS June--August 1995

    SciTech Connect

    Arfelli, F.; Burns, C.; Chapman, D. |

    1995-12-31

    At Brookhaven National Laboratory mammography experiments are being carried out at the X27C R and D beamline of the National Synchrotron Light Source using a monochromatic x-ray beam in order to explore the potential of monoenergetic photons for mammographic imaging. In two different periods of beamtime the authors have performed preliminary studies of mammographic imaging using a monochromatic synchrotron radiation source. They used both phantom objects and real tissue samples. Qualitative studies with the contrast-detail phantom show good agreement when compared with the theoretical contrast. As expected, the contrast is higher if the energy is lower. The results show an improved contrast with energies 18 keV and lower compared to images obtained from conventional polyenergetic x-ray imaging systems. The results also show that for similar imaging conditions the monoenergetic mean glandular dose is less than that from polyenergetic sources. This is due both to the increased sensitivity of the image plate detectors and to actual reductions of dose for truly monochromatic beams.

  8. Anatomical noise in contrast-enhanced digital mammography. Part II. Dual-energy imaging

    SciTech Connect

    Hill, Melissa L.; Yaffe, Martin J.; Mainprize, James G.; Carton, Ann-Katherine; Saab-Puong, Sylvie; Iordache, R?zvan; Muller, Serge; Jong, Roberta A.; Dromain, Clarisse

    2013-08-15

    Purpose: Dual-energy (DE) contrast-enhanced digital mammography (CEDM) uses an iodinated contrast agent in combination with digital mammography (DM) to evaluate lesions on the basis of tumor angiogenesis. In DE imaging, low-energy (LE) and high-energy (HE) images are acquired after contrast administration and their logarithms are subtracted to cancel the appearance of normal breast tissue. Often there is incomplete signal cancellation in the subtracted images, creating a background “clutter” that can impair lesion detection. This is the second component of a two-part report on anatomical noise in CEDM. In Part I the authors characterized the anatomical noise for single-energy (SE) temporal subtraction CEDM by a power law, with model parameters ? and ?. In this work the authors quantify the anatomical noise in DE CEDM clinical images and compare this with the noise in SE CEDM. The influence on the anatomical noise of the presence of iodine in the breast, the timing of imaging postcontrast administration, and the x-ray energy used for acquisition are each evaluated.Methods: The power law parameters, ? and ?, were measured from unprocessed LE and HE images and from DE subtracted images to quantify the anatomical noise. A total of 98 DE CEDM cases acquired in a previous clinical pilot study were assessed. Conventional DM images from 75 of the women were evaluated for comparison with DE CEDM. The influence of the imaging technique on anatomical noise was determined from an analysis of differences between the power law parameters as measured in DM, LE, HE, and DE subtracted images for each subject.Results: In DE CEDM, weighted image subtraction lowers ? to about 1.1 from 3.2 and 3.1 in LE and HE unprocessed images, respectively. The presence of iodine has a small but significant effect in LE images, reducing ? by about 0.07 compared to DM, with ? unchanged. Increasing the x-ray energy, from that typical in DM to a HE beam, significantly decreases ? by about 2 × 10{sup ?5} mm{sup 2}, and lowers ? by about 0.14 compared to LE images. A comparison of SE and DE CEDM at 4 min postcontrast shows equivalent power law parameters in unprocessed images, and lower ? and ? by about 3 × 10{sup ?5} mm{sup 2} and 0.50, respectively, in DE versus SE subtracted images.Conclusions: Image subtraction in both SE and DE CEDM reduces ? by over a factor of 2, while maintaining ? below that in DM. Given the equivalent ? between SE and DE unprocessed CEDM images, and the smaller anatomical noise in the DE subtracted images, the DE approach may have an advantage over SE CEDM. It will be necessary to test this potential advantage in future lesion detectability experiments, which account for realistic lesion signals. The authors' results suggest that LE images could be used in place of DM images in CEDM exam interpretation.

  9. Applying the European protocol for the quality control of the physical and technical aspects of mammography screening threshold contrast visibility assessment to digital systems

    NASA Astrophysics Data System (ADS)

    Van Metter, Richard; Heath, Michael; Fletcher-Heath, Lynn

    2006-03-01

    The need to assure the image quality of digital systems for mammography screening applications is now widely recognized. One approach is embodied in Part B of the European Protocol for the Quality Control of the Physical and Technical Aspects of Mammography Screening (EPQCM), which prescribes criteria for several interconnected image quality metrics. The focus of this study is on the "threshold contrast visibility" (TCV) protocol (section 2.4.1 of the EPQCM), in which human observers score images of a CDMAM or similar 4-AFC phantom. This section of the EPQCM currently omits many critical experimental details, which must be gleaned from ancillary documents. Given these, the purpose of this study is to quantify the effects of several remaining experimental variables, including phantom design, and the methods used for scoring and analysis, on the measured results. Preliminary studies of two CDMAM version 3.4 (CDMAM 3.4) phantoms have revealed a 17% difference in TCV when averaged over all target diameters from 0.1 to 2.0 mm. This indicates phantom variability may affect results at some sites. More importantly, we have shown that the current CDMAM phantom design, methods for scoring, and analysis, substantially limit the ability to measure system performance accurately and precisely. An improved phantom design has been shown to avoid these limitations. Viewing environment and presentation context affect the performance and efficiency of visual scoring of phantom images. An automated display tool has been developed that isolates individual 4-AFC targets of CDMAM phantom images, automatically optimizes window/level, and automatically records observers' scores. While not substantially changing TCV, the tool has increased scoring efficiency while mitigating several of the limitations associated with unassisted visual scoring. For example, learning bias and navigational issues are completely avoided. Ultimately, software-based ideal observer scoring will likely prove to be a better approach. Statistical-decision-theory-based (SDT) analysis has been shown to mitigate limitations associated with the current CDMAM phantom and the ad hoc nearest-neighbor correcting (NNC) scoring method. NNC analysis is sensitive to the degree of incomplete scoring (stopping criteria). However, SDT substantially mitigates this problem, using all of the available data to derive thresholds that are more interpretable. Bootstrap sampling was used to provide an estimate of the standard error for SDT analysis. In conclusion, the current EPQCM section 2.4.1 protocol fails to measure TCV accurately and precisely enough to qualify digital mammography systems. This paper presents a series of recommendations that supplement section 2.4.1 of the EPQCM and that provide a stable and accurate measure of TCV.

  10. Anatomical noise in contrast-enhanced digital mammography. Part I. Single-energy imaging

    SciTech Connect

    Hill, Melissa L.; Yaffe, Martin J.; Mainprize, James G.; Carton, Ann-Katherine; Muller, Serge; Ebrahimi, Mehran; Jong, Roberta A.; Dromain, Clarisse

    2013-05-15

    Purpose: The use of an intravenously injected iodinated contrast agent could help increase the sensitivity of digital mammography by adding information on tumor angiogenesis. Two approaches have been made for clinical implementation of contrast-enhanced digital mammography (CEDM), namely, single-energy (SE) and dual-energy (DE) imaging. In each technique, pairs of mammograms are acquired, which are then subtracted with the intent to cancel the appearance of healthy breast tissue to permit sensitive detection and specific characterization of lesions. Patterns of contrast agent uptake in the healthy parenchyma, and uncanceled signal from background tissue create a 'clutter' that can mask or mimic an enhancing lesion. This type of 'anatomical noise' is often the limiting factor in lesion detection tasks, and thus, noise quantification may be useful for cascaded systems analysis of CEDM and for phantom development. In this work, the authors characterize the anatomical noise in CEDM clinical images and the authors evaluate the influence of the x-ray energy used for acquisition, the presence of iodine in the breast, and the timing of imaging postcontrast administration on anatomical noise. The results are presented in a two-part report, with SE CEDM described here, and DE CEDM in Part II. Methods: A power law is used to model anatomical noise in CEDM images. The exponent, {beta}, which describes the anatomical structure, and the constant {alpha}, which represents the magnitude of the noise, are determined from Wiener spectra (WS) measurements on images. A total of 42 SE CEDM cases from two previous clinical pilot studies are assessed. The parameters {alpha} and {beta} are measured both from unprocessed images and from subtracted images. Results: Consistent results were found between the two SE CEDM pilot studies, where a significant decrease in {beta} from a value of approximately 3.1 in the unprocessed images to between about 1.1 and 1.8 in the subtracted images was observed. Increasing the x-ray energy from that used in conventional DM to those of typical SE CEDM spectra with mean energies above 33 keV significantly decreased {alpha} by about a factor of 19, in agreement with theory. Compared to precontrast images, in the unprocessed postcontrast images at 30 s postinjection, {alpha} was larger by about 7.4 Multiplication-Sign 10{sup -7} mm{sup 2} and {beta} was decreased by 0.2. While {alpha} did not vary significantly with the time after contrast administration, {beta} from the unprocessed image WS increased linearly, and {beta} from subtracted image WS increased with an initial quadratic relationship that plateaued by about 5 min postinjection. Conclusions: The presence of an iodinated contrast agent in the breast produced small, but significant changes in the power law parameters of unprocessed CEDM images compared to the precontrast images. Image subtraction in SE CEDM significantly reduced anatomical noise compared to conventional DM, with a reduction in both {alpha} and {beta} by about a factor of 2. The data presented here, and in Part II of this work, will be useful for modeling of CEDM backgrounds, for systems characterization and for lesion detectability experiments using models that account for anatomical noise.

  11. A comparison of image interpretation times in full field digital mammography and digital breast tomosynthesis

    NASA Astrophysics Data System (ADS)

    Astley, Susan; Connor, Sophie; Lim, Yit; Tate, Catriona; Entwistle, Helen; Morris, Julie; Whiteside, Sigrid; Sergeant, Jamie; Wilson, Mary; Beetles, Ursula; Boggis, Caroline; Gilbert, Fiona

    2013-03-01

    Digital Breast Tomosynthesis (DBT) provides three-dimensional images of the breast that enable radiologists to discern whether densities are due to overlapping structures or lesions. To aid assessment of the cost-effectiveness of DBT for screening, we have compared the time taken to interpret DBT images and the corresponding two-dimensional Full Field Digital Mammography (FFDM) images. Four Consultant Radiologists experienced in reading FFDM images (4 years 8 months to 8 years) with training in DBT interpretation but more limited experience (137-407 cases in the past 6 months) were timed reading between 24 and 32 two view FFDM and DBT cases. The images were of women recalled from screening for further assessment and women under surveillance because of a family history of breast cancer. FFDM images were read before DBT, according to local practice. The median time for readers to interpret FFDM images was 17.0 seconds, with an interquartile range of 12.3-23.6 seconds. For DBT, the median time was 66.0 seconds, and the interquartile range was 51.1-80.5 seconds. The difference was statistically significant (p<0.001). Reading times were significantly longer in family history clinics (p<0.01). Although it took approximately four times as long to interpret DBT than FFDM images, the cases were more complex than would be expected for routine screening, and with higher mammographic density. The readers were relatively inexperienced in DBT interpretation and may increase their speed over time. The difference in times between clinics may be due to increased throughput at assessment, or decreased density.

  12. Diagnostic clinical benefits of digital spot and digital 3D mammography following analysis of screening findings

    NASA Astrophysics Data System (ADS)

    Lehtimaki, Mari; Pamilo, Martti; Raulisto, Leena; Roiha, Marja; Kalke, Martti; Siltanen, Samuli; Ihamäki, Timo

    2003-05-01

    The purpose of this study is to find out the impact of 3-dimensional digital mammography and digital spot imaging following analysis of the abnormal findings of screening mammograms. Over a period of eight months, digital 3-D mammography imaging TACT Tuned Aperture Computed Tomography+, digital spot imaging (DSI), screen-film mammography imaging (SFM) and diagnostic film imaging (DFM) examinations were performed on 60 symptomatic cases. All patients were recalled because it was not possible to exclude the presence of breast cancer on screening films. Abnormal findings on the screening films were non-specific tumor-like parenchymal densities, parenchymal asymmetries or distortions with or without microcalcifications or just microcalcifications. Mammography work-up (film imaging) included spot compression and microfocus magnification views. The 3-D softcopy reading in all cases was done with Delta 32 TACT mammography workstation, while the film images were read using a mammography-specific light box. During the softcopy reading only windowing tools were allowed. The result of this study indicates that the clinical diagnostic image quality of digital 3-D and digital spot images are better than in film images, even in comparison with diagnostic work-up films. Potential advantages are to define if the mammography finding is caused by a real abnormal lesion or by superimposition of normal parenchymal structures, to detect changes in breast tissue which would otherwise be missed, to verify the correct target for biopsies and to reduce the number of biopsies performed.

  13. First results with real-time selenium-based full-field digital mammography three-dimensional imaging system

    NASA Astrophysics Data System (ADS)

    Lehtimaki, Mari; Pamilo, Martti; Raulisto, Leena; Kalke, Martti

    2004-05-01

    Our goal in this paper is to evaluate the capability of real-time selenium-technology-based full-field digital mammography (FFDM) system in breast tomosynthesis. The objective of this study is to find out the present status of amorphous selenium technology in the sense of advanced applications in clinical use. We were using tuned aperture computed tomography (TACT+) 3-dimensional (3D) technology for reconstruction. Under evaluation were amorphous selenium signal-to-noise-ratio, flat panel image artefacts and acquisition time to perform full-field digital mammography 3D examination. To be able to validate the system we used a special breast phantom. We found out that 3D imaging technology provides diagnostic value and benefits over 2-dimensional (2D) imaging. 3D TACT advantages are to define if mammography finding is caused by a real abnormal lesion or by superposition of normal parenchymal structures, to be able to diagnose and analyze the findings properly, to detect changes in breast tissue which would otherwise be missed, to verify the possible multifocality of the breast cancers, to verify the correct target for biopsies and to reduce number of biopsies performed. Slice visualization and 3D volume model provide greater diagnostic information compared to 2D projection screening and diagnostic imaging.

  14. Potential missed detection with screening mammography: does the quality of radiologist’s interpretation vary by patient socioeconomic advantage/disadvantage

    PubMed Central

    Rauscher, Garth H; Khan, Jenna A; Berbaum, Michael L; Conant, Emily F

    2013-01-01

    PURPOSE We examined whether quality of mammography interpretation as performed by the original reading radiologist varied by patient sociodemographic characteristics. METHODS For 149 patients residing in Chicago and diagnosed in 2005-2008, we obtained the original index mammogram that detected the breast cancer and at least one prior mammogram that did not detect the cancer performed within 2 years of the index mammogram. A single breast imaging specialist performed a blinded review of the prior mammogram. Potentially missed detection was defined as an actionable lesion seen during a blinded review of the prior mammogram that was in the same quadrant as the cancer on the index mammogram. RESULTS Of 149 prior mammograms originally read as non-malignant, 46% (N=68) had a potentially detectable lesion. In unadjusted analyses, potentially missed detection was greater among minority patients (54% vs. 39%, p=0.07), for patients with incomes below $30,000 (65% vs. 36%, p<0.01), with less education (58% vs. 39%, p=0.02), and lacking private health insurance (63% vs. 40%, p=0.02). Likelihood ratio tests for the inclusion of socioeconomic variables in multivariable logistic regression models were highly significant (p<=0.02). CONCLUSIONS Disadvantaged socioeconomic status appears to be associated with potentially missed detection of breast cancer at mammography screening. PMID:23453384

  15. Using LROC analysis to evaluate detection accuracy of microcalcification clusters imaged with flat-panel CT mammography

    NASA Astrophysics Data System (ADS)

    Gong, Xing; Glick, Stephen J.; Vedula, Aruna A.

    2004-05-01

    The purpose of this study is to investigate the detectability of microcalcification clusters (MCCs) using CT mammography with a flat-panel detector. Compared with conventional mammography, CT mammography can provide improved discrimination between malignant and benign cases as it can provide the radiologist with more accurate morphological information on MCCs. In this study, two aspects of MCC detection with flat-panel CT mammography were examined: (1) the minimal size of MCCs detectable with mean glandular dose (MGD) used in conventional mammography; (2) the effect of different detector pixel size on the detectability of MCCs. A realistic computer simulation modeling x-ray transport through the breast, as well as both signal and noise propagation through the flat-panel imager, was developed to investigate these questions. Microcalcifications were simulated as calcium carbonate spheres with diameters set at the levels of 125, 150 and 175 ?m. Each cluster consisted of 10 spheres spread randomly in a 6×6 mm2 region of interest (ROI) and the detector pixel size was set to 100×100, 200×200, or 300×300?m2. After reconstructing 100 projection sets for each case (half with signal present) with the cone-beam Feldkamp (FDK) algorithm, a localization receiver operating characteristic (LROC) study was conducted to evaluate the detectability of MCCs. Five observers chose the locations of cluster centers with correspondent confidence ratings. The average area under the LROC curve suggested that the 175 ?m MCCs can be detected at a high level of confidence. Results also indicate that flat-panel detectors with pixel size of 200×200 ?m2 are appropriate for detecting small targets, such as MCCs.

  16. NPS comparison of anatomical noise characteristics in mammography, tomosynthesis, and breast CT images using power law metrics

    NASA Astrophysics Data System (ADS)

    Chen, Lin; Boone, John M.; Nosratieh, Anita; Abbey, Craig K.

    2011-03-01

    Digital mammography is the current standard for breast cancer screening, however breast tomosynthesis and breast CT (bCT) have been studied in clinical trials. At our institution, 30 women (BIRADS 4 and 5) underwent IRB-approved imaging by mammography, breast tomosynthesis, and bCT on the same day. Twenty three data sets were used for analysis. The 2D noise power spectrum (NPS) was computed and averaged for each data set. The NPS was computed for different slice thicknesses of dx × N, where dx ~ 0.3 mm and N=1-64, on the bCT data. Each 2D NPS was radially averaged, and the 1D data were fit using a power law function as proposed by Burgess: NPS(f) = αf-β. The value of β was determined over a range of frequencies corresponding to anatomical noise, for each patient and each modality. Averaged over the 30 women (26 for bCT, 28 for tomosynthesis, 28 for mammography), for mammography β=3.06 (0.25), for CC tomosynthesis β=2.91 (0.35), and for axial bCT β=1.72 (0.47). For sagittal bCT β=1.77 (0.36) and for coronal bCT, β=1.88 (0.45). The computation of β versus slice thickness on the coronal bCT data set led to β~1.7 for N=1, asymptotically reaching β ~ 3 for larger slice thickness. These results suggest that there is a fundamental difference in breast anatomic noise as characterized by β, between thin slices (<2 mm) and thicker slices. Tomosynthesis was found to have anatomic noise properties closer to mammography than breast CT, most likely due to the relatively thick slice sensitivity profile of tomosynthesis.

  17. A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone-beam CT breast imaging

    SciTech Connect

    Gong Xing; Glick, Stephen J.; Liu, Bob; Vedula, Aruna A.; Thacker, Samta

    2006-04-15

    Although conventional mammography is currently the best modality to detect early breast cancer, it is limited in that the recorded image represents the superposition of a three-dimensional (3D) object onto a 2D plane. Recently, two promising approaches for 3D volumetric breast imaging have been proposed, breast tomosynthesis (BT) and CT breast imaging (CTBI). To investigate possible improvements in lesion detection accuracy with either breast tomosynthesis or CT breast imaging as compared to digital mammography (DM), a computer simulation study was conducted using simulated lesions embedded into a structured 3D breast model. The computer simulation realistically modeled x-ray transport through a breast model, as well as the signal and noise propagation through a CsI based flat-panel imager. Polyenergetic x-ray spectra of Mo/Mo 28 kVp for digital mammography, Mo/Rh 28 kVp for BT, and W/Ce 50 kVp for CTBI were modeled. For the CTBI simulation, the intensity of the x-ray spectra for each projection view was determined so as to provide a total average glandular dose of 4 mGy, which is approximately equivalent to that given in conventional two-view screening mammography. The same total dose was modeled for both the DM and BT simulations. Irregular lesions were simulated by using a stochastic growth algorithm providing lesions with an effective diameter of 5 mm. Breast tissue was simulated by generating an ensemble of backgrounds with a power law spectrum, with the composition of 50% fibroglandular and 50% adipose tissue. To evaluate lesion detection accuracy, a receiver operating characteristic (ROC) study was performed with five observers reading an ensemble of images for each case. The average area under the ROC curves (A{sub z}) was 0.76 for DM, 0.93 for BT, and 0.94 for CTBI. Results indicated that for the same dose, a 5 mm lesion embedded in a structured breast phantom was detected by the two volumetric breast imaging systems, BT and CTBI, with statistically significant higher confidence than with planar digital mammography, while the difference in lesion detection between BT and CTBI was not statistically significant.

  18. A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone-beam CT breast imaging.

    PubMed

    Gong, Xing; Glick, Stephen J; Liu, Bob; Vedula, Aruna A; Thacker, Samta

    2006-04-01

    Although conventional mammography is currently the best modality to detect early breast cancer, it is limited in that the recorded image represents the superposition of a three-dimensional (3D) object onto a 2D plane. Recently, two promising approaches for 3D volumetric breast imaging have been proposed, breast tomosynthesis (BT) and CT breast imaging (CTBI). To investigate possible improvements in lesion detection accuracy with either breast tomosynthesis or CT breast imaging as compared to digital mammography (DM), a computer simulation study was conducted using simulated lesions embedded into a structured 3D breast model. The computer simulation realistically modeled x-ray transport through a breast model, as well as the signal and noise propagation through a CsI based flat-panel imager. Polyenergetic x-ray spectra of Mo/Mo 28 kVp for digital mammography, Mo/Rh 28 kVp for BT, and W/Ce 50 kVp for CTBI were modeled. For the CTBI simulation, the intensity of the x-ray spectra for each projection view was determined so as to provide a total average glandular dose of 4 mGy, which is approximately equivalent to that given in conventional two-view screening mammography. The same total dose was modeled for both the DM and BT simulations. Irregular lesions were simulated by using a stochastic growth algorithm providing lesions with an effective diameter of 5 mm. Breast tissue was simulated by generating an ensemble of backgrounds with a power law spectrum, with the composition of 50% fibroglandular and 50% adipose tissue. To evaluate lesion detection accuracy, a receiver operating characteristic (ROC) study was performed with five observers reading an ensemble of images for each case. The average area under the ROC curves (Az) was 0.76 for DM, 0.93 for BT, and 0.94 for CTBI. Results indicated that for the same dose, a 5 mm lesion embedded in a structured breast phantom was detected by the two volumetric breast imaging systems, BT and CTBI, with statistically significant higher confidence than with planar digital mammography, while the difference in lesion detection between BT and CTBI was not statistically significant. PMID:16696481

  19. Dual-energy digital mammography for calcification imaging: theory and implementation

    NASA Astrophysics Data System (ADS)

    Kappadath, Srinivas C.; Shaw, Chris C.; Lai, Chao-Jen; Liu, Xinming; Whitman, Gary

    2004-05-01

    Small microcalcifications essential to the early detection of breast cancer may be obscured by overlapping tissue structures. Dual-energy digital mammography (DEDM), where separate low- and high-energy images are acquired and synthesized to cancel the tissue structures, may improve the ability to detect and visualize microcalcifications. The investigation of DEDM began with a signal-to noise ratio analysis to estimate and relate the noise level in the dual-energy calcification signals to the x-ray spectra, microcalcification size, tissue composition and breast thickness. We investigated various inverse-mapping functions, both linear and non-linear, to estimate the calcification thickness from low- and high-energy measurements. Transmission (calibration) measurements made at two different kVp values for variable aluminum thickness (to simulate calcifications) and variable glandular-tissue ratio for a fixed total tissue thickness were used to determine the coefficients of the inverse-mapping functions by a least-squares analysis. We implemented and evaluated the DEDM technique under narrow-beam geometry. Phantoms, used in the evaluation, were constructed by placing different aluminum strips over breast-tissue-equivalent materials of different compositions. The resulting phantom images consisted of four distinct regions, each with a different combination of aluminum thickness and tissue composition. DEDM with non-linear inverse-mapping functions could successfully cancel the contrast of the tissue-structure background to better visualize the overlapping aluminum strip. We are currently in the process of translating our DEDM techniques into full-field imaging. We have designed special phantoms with variable glandular ratios and variable calcification thicknesses for evaluation of the full-field dual-energy calcification images.

  20. Performance Benchmarks for Diagnostic Mammography

    Cancer.gov

    In the United States, a Food and Drug Administration (FDA) regulation requires limited auditing of clinical outcomes for all screening and diagnostic mammography examinations that have been assessed as either suspicious for malignancy or highly suggestive of malignancy. More comprehensive auditing is performed by many mammography facilities in both the United States and other countries. Auditing is thought to be a useful quality assurance procedure, providing performance feedback to both mammography facilities and individual interpreting radiologists.

  1. Visibility of microcalcification clusters and masses in breast tomosynthesis image volumes and digital mammography: A 4AFC human observer study

    SciTech Connect

    Timberg, P.; Baath, M.; Andersson, I.; Mattsson, S.; Tingberg, A.; Ruschin, M.

    2012-05-15

    Purpose: To investigate the visibility of simulated lesions in digital breast tomosynthesis (BT) image volumes compared with 2D digital mammography (DM). Methods: Simulated lesions (masses and microcalcifications) were added to images of the same women acquired on a DM system (Mammomat Novation, Siemens) and a BT prototype. The same beam quality was used for the DM and BT acquisitions. The total absorbed dose resulting from a 25-projection BT acquisition and reconstruction (BT{sub 25}) was approximately twice that of a single DM view. By excluding every other projection image from the reconstruction (BT{sub 13}), approximately the same dose as in DM was effected. Simulated microcalcifications were digitally added with varying contrast to the DM and BT images. Simulated masses with 8 mm diameter were also added to BT images. A series of 4-alternative forced choice (4AFC) human observer experiments were conducted. Four medical physicists participated in all experiments, each consisting of 60 trials per experimental condition. The observers interpreted the BT image volumes in cine-mode at a fixed image sequence speed. The required threshold contrast (S{sub t}) to achieve a detectability index (d') of 2.5 (i.e., 92.5% correct decisions) was determined. Results: The S{sub t} for mass detection in DM was approximately a factor of 2 higher than required in BT indicating that the detection of masses was improved under BT conditions compared to DM. S{sub t} for microcalcification detection was higher for BT than for DM at both BT dose levels (BT{sub 25} and BT{sub 13}), with a statistically significant difference in S{sub t} between DM and BT{sub 13}. These results indicate a dose-dependent decrease in detection performance in BT for detection of microcalcifications. Conclusions: In agreement with previous investigations, masses of size 8 mm can be detected with less contrast in BT than in DM indicating improved detection performance for BT. However, for the investigated microcalcifications, the results of this study indicate potentially worse performance for BT than for DM at the same dose level.

  2. EDITORIAL: Optical mammography: Imaging and characterization of breast lesions by pulsed near-infrared laser light (OPTIMAMM)

    NASA Astrophysics Data System (ADS)

    Hebden, Jeremy C.; Rinneberg, Herbert

    2005-06-01

    The Commission of the European Union (EU) conceived its Fifth Framework Programme (FP5) to identify the priorities for the European Union's research, technological development and demonstration activities for the period 1998-2002. By encouraging collaborative research between groups in different member countries, FP5 was intended to help solve problems the EU is facing and respond to major socio-economic challenges. The programme focused on a number of objectives and areas combining technological, industrial, economic, social and cultural aspects. A specific call was made, under its `Quality of Life and Management of Living Resources' section, for proposals which aim to explore improvements in non-invasive methods of imaging for early diagnosis and clinical evaluation of disease. Among the projects successfully funded under the FP5 programme was one entitled `Optical mammography: Imaging and characterization of breast lesions by pulsed near-infrared laser light', known by its acronym OPTIMAMM. The project involved a consortium of nine partners, comprising ten applied science and clinical research groups based in six EU countries, with overall administration and management provided by the Physikalisch-Technische Bundesanstalt, Berlin, Germany. The broad aim of the OPTIMAMM project was to combine multi-disciplinary basic (physics, engineering, mathematics, computer science) and clinical (oncology, histology) research to assess the diagnostic potential of time-domain optical and photoacoustic mammography as novel, non-invasive imaging modalities for the detection and clinical evaluation of breast lesions. Funding for the project, at a total cost of about 1.67 MEuro, began in December 2000 for a period of three years, although a zero-cost extension was granted to enable the ongoing project activities to continue until the end of May 2004. The importance of developing new tools for the detection and diagnosis of breast disease is evident from the very high incidence and mortality associated with it, within the EU and throughout the world. Although x-ray mammography is recognized as an effective tool for cancer screening in women over 35-40 years of age, it suffers from a significant number of false positives which often lead to unnecessary biopsy. X-ray mammography is also less effective for younger women with denser breasts, and involves the use of potentially harmful ionizing radiation. While other conventional diagnostic techniques such as ultrasound and magnetic resonance imaging (MRI) are also widely used in the diagnosis and characterization of breast disease, their roles in the detection and staging of breast tumours have so far been limited. The development of optical methods of imaging the breast is attractive partly because they are safe, but chiefly because they can reveal contrast between normal and diseased tissues which are not evident using conventional methods. The principal mechanism for contrast at near-infrared wavelengths is the characteristic absorption by haemoglobin and other dominant tissue chromophores, such as fat and water. Furthermore, the differences between the absorption of oxy-haemoglobin and deoxy-hemoglobin provide a means of determining oxygenation, and therefore of studying tissue function. The OPTIMAMM project focused specifically on the diagnostic potential of time-resolved methods. Systems which measure the flight-times of photons transmitted across highly scattering breast tissue offer the potential to provide greater spatial resolution and contrast than systems based on intensity measurements alone, and facilitate better separation between the effects of scatter and those of absorption. A major component of the project was a series of clinical trials performed at four European sites, in particular in Berlin (Germany) and Milan (Italy) using similar scanning instrumentation, carried out under a harmonized clinical protocol where appropriate. The clinical trials were augmented by efforts to refine semi-empirical and rigorous mathematical methods for data analysis and image reconstruction, and by technical improvements to instrumentation. Throughout the project, developments in technology and methodology were assessed through appropriate evaluation on human subjects. The nine project partners also applied different measurement techniques to a concerted effort to reliably measure the scattering and absorption properties of normal breast tissue, benign lesions and tumours at various near-infrared wavelengths in vivo non-invasively, during breast surgery, and ex vivo on tissue specimens and biopsy samples. In May 2004, on the successful completion of the OPTIMAMM project, a European workshop entitled `Applied Medical Photonics: from Tissue Characterization to Optical Mammography' was held in Berlin, jointly organized by the OPTIMAMM consortium and by another consortium representing the European network on `Medical Photonics'. At this meeting, all the the OPTIMAMM partners presented summaries of their results obtained during the lifetime of the project. Subsequently, the ten collaborating research groups have prepared a series of papers which focus on the specific results of the consortia and review its overall achievements. This special issue of Physics in Medicine and Biology presents these ten papers (pages 2429-2596), which have been accepted for publication following the usual thorough peer-review process for this journal. The first four papers describe the results of the two parallel clinical assessments of scanning time-domain optical mammography, involving a combined total of about 350 patients. These include descriptions of technical developments and the results of efforts to determine the optical properties of benign lesions and carcinomas. The first two contributions summarize the work performed by collaborating groups based in Berlin. A scanning laser pulse optical mammograph, which records craniocaudal and mediolateral projection images of the compressed breast at three near-infrared wavelengths, was used in a clinical trial on 154 patients, all undergoing subsequent histological assessment of at least one breast lesion. Part I describes an analysis of optical images by comparing them with x-ray and MRI mammograms and with results of histopathology. Part II reports on the derived optical properties of a subset of 87 carcinomas. Haemoglobin concentration was found to be systematically higher in tumours than in normal tissue, while no statistical difference was observed for blood oxygenation. A similar breast imaging system is the focus of the next two papers, presented by partners based in Milan. A clinical study involving a total of 194 patients was performed over the lifetime of the project, during which the number of discrete wavelengths employed increased from four to seven, and the spectral range was expanded to accommodate wavelengths above 900 nm. The first of this pair of papers describes the effectiveness of the system for detecting breast lesions, while the second presents an attempt to characterize the optical properties of malignant and benign lesions. Again, cancers were found to associate with a blood content that is higher than in surrounding tissues. The fifth paper in this special issue, authored by consortium partners in London (UK), reports clinical results achieved using a 32-channel time-resolved optical tomography system. Cross-sectional images of the uncompressed human breast are generated from photon flight-time measurements using an iterative, non-linear algorithm. Studies involving patients with a variety of breast conditions detected 17 out of 19 lesions, and results on healthy volunteers displayed heterogeneity which was repeatable over a period of months. The next paper represents a collaboration between the London researchers working on image reconstruction algorithm development, and partners based in Berlin. A paraxial scanning mechanism developed for the Berlin breast imaging system was used to acquire measurements which enable three-dimensional (3D) in vivo images to be generated. A fast reconstruction method, based on the Rytov approximation, uses data which is Fourier transformed with respect to both time and space. The seventh paper, contributed by partners based in Enschede (The Netherlands), provides an overview of their breast imaging system which measures photoacoustic signals occurring within tissue in response to illumination by short pulses of light. The overall performance of the system is discussed, and an algorithm is presented which is designed to utilize these measurements to generate 3D images of the compressed breast. The remaining three papers focus on the measurement of the optical properties of breast tissue. The first of these presents the results of an investigation of the intra- and inter-subject variability of optical properties, and of the relative concentrations of principle chromophores: water, lipid, and haemoglobin. This collaborative study, involving partners based in Lund (Sweden) and Milan, utilized in vivo time-resolved spectroscopy measurements at four near-infrared wavelengths. Differences between pre- and post-menopausal breast tissues were confirmed, and no systematic differences between contralateral breast were evident. The next paper, authored by researchers in Rotterdam (The Netherlands), describes the use of a fibre-optic needle probe and a technique known as differential pathlength spectroscopy to derive highly localized optical properties of the breast in vivo. Results, correlated with histological outcome, reveal that malignant tissue is characterized by a significant decrease in tissue oxygenation and a higher blood content compared with normal breast tissue. The last paper in this special issue describes a theoretical development which enables a more accurate determination of scattering properties from small biopsy samples. The theory was applied by the authors based in Heraklion (Greece) to time-resolved transmittance measurements in order to characterize the scattering of 59 samples of breast tissue. We are grateful to all the authors for their valuable contributions and their prompt responses to reviewers' comments, and we thank all the reviewers for their useful suggestions to improve the quality of the papers and their considerable efforts to meet tight deadlines. And last but not least we would like to thank the European Commission for its generous financial support.

  3. A Novel Method to Assess Incompleteness of Mammography Reports

    PubMed Central

    Gimenez, Francisco J.; Wu, Yirong; Burnside, Elizabeth S.; Rubin, Daniel L.

    2014-01-01

    Mammography has been shown to improve outcomes of women with breast cancer, but it is subject to inter-reader variability. One well-documented source of such variability is in the content of mammography reports. The mammography report is of crucial importance, since it documents the radiologist’s imaging observations, interpretation of those observations in terms of likelihood of malignancy, and suggested patient management. In this paper, we define an incompleteness score to measure how incomplete the information content is in the mammography report and provide an algorithm to calculate this metric. We then show that the incompleteness score can be used to predict errors in interpretation. This method has 82.6% accuracy at predicting errors in interpretation and can possibly reduce total diagnostic errors by up to 21.7%. Such a method can easily be modified to suit other domains that depend on quality reporting. PMID:25954448

  4. A novel method to assess incompleteness of mammography reports.

    PubMed

    Gimenez, Francisco J; Wu, Yirong; Burnside, Elizabeth S; Rubin, Daniel L

    2014-01-01

    Mammography has been shown to improve outcomes of women with breast cancer, but it is subject to inter-reader variability. One well-documented source of such variability is in the content of mammography reports. The mammography report is of crucial importance, since it documents the radiologist's imaging observations, interpretation of those observations in terms of likelihood of malignancy, and suggested patient management. In this paper, we define an incompleteness score to measure how incomplete the information content is in the mammography report and provide an algorithm to calculate this metric. We then show that the incompleteness score can be used to predict errors in interpretation. This method has 82.6% accuracy at predicting errors in interpretation and can possibly reduce total diagnostic errors by up to 21.7%. Such a method can easily be modified to suit other domains that depend on quality reporting. PMID:25954448

  5. Optimization of breast cancer detection in Dual Energy X-ray Mammography using a CMOS imaging detector

    NASA Astrophysics Data System (ADS)

    Koukou, V.; Fountos, G.; Martini, N.; Sotiropoulou, P.; Michail, C.; Kalyvas, N.; Valais, I.; Bakas, A.; Kounadi, E.; Kandarakis, I.; Nikiforidis, G.

    2015-01-01

    Dual energy mammography has the ability to improve the detection of microcalcifications leading to early diagnosis of breast cancer. In this simulation study, a prototype dual energy mammography system, using a CMOS based imaging detector with different X-ray spectra, was modeled. The device consists of a 33.91 mg/cm2 Gd2O2S:Tb scintillator screen, placed in direct contact with the sensor, with a pixel size of 22.5 μm. Various filter materials and tube voltages of a Tungsten (W) anode for both the low and high energy were examined. The selection of the filters applied to W spectra was based on their K- edges (K-edge filtering). Hydroxyapatite (HAp) was used to simulate microcalcifications. Calcification signal-to-noise ratio (SNRtc) was calculated for entrance surface dose within the acceptable levels of conventional mammography. Optimization was based on the maximization of SNRtc while minimizing the entrance dose. The best compromise between SNRtc value and dose was provided by a 35kVp X-ray spectrum with added beam filtration of 100μm Pd and a 70kVp Yb filtered spectrum of 800 μm for the low and high energy, respectively. Computer simulation results show that a SNRtc value of 3.6 can be achieved for a calcification size of 200 μm. Compared with previous studies, this method can improve detectability of microcalcifications.

  6. Comparison of breast tissue measurements using magnetic resonance imaging, digital mammography and a mathematical algorithm

    NASA Astrophysics Data System (ADS)

    Lu, Lee-Jane W.; Nishino, Thomas K.; Johnson, Raleigh F.; Nayeem, Fatima; Brunder, Donald G.; Ju, Hyunsu; Leonard, Morton H., Jr.; Grady, James J.; Khamapirad, Tuenchit

    2012-11-01

    Women with mostly mammographically dense fibroglandular tissue (breast density, BD) have a four- to six-fold increased risk for breast cancer compared to women with little BD. BD is most frequently estimated from two-dimensional (2D) views of mammograms by a histogram segmentation approach (HSM) and more recently by a mathematical algorithm consisting of mammographic imaging parameters (MATH). Two non-invasive clinical magnetic resonance imaging (MRI) protocols: 3D gradient-echo (3DGRE) and short tau inversion recovery (STIR) were modified for 3D volumetric reconstruction of the breast for measuring fatty and fibroglandular tissue volumes by a Gaussian-distribution curve-fitting algorithm. Replicate breast exams (N = 2 to 7 replicates in six women) by 3DGRE and STIR were highly reproducible for all tissue-volume estimates (coefficients of variation <5%). Reliability studies compared measurements from four methods, 3DGRE, STIR, HSM, and MATH (N = 95 women) by linear regression and intra-class correlation (ICC) analyses. Rsqr, regression slopes, and ICC, respectively, were (1) 0.76-0.86, 0.8-1.1, and 0.87-0.92 for %-gland tissue, (2) 0.72-0.82, 0.64-0.96, and 0.77-0.91, for glandular volume, (3) 0.87-0.98, 0.94-1.07, and 0.89-0.99, for fat volume, and (4) 0.89-0.98, 0.94-1.00, and 0.89-0.98, for total breast volume. For all values estimated, the correlation was stronger for comparisons between the two MRI than between each MRI versus mammography, and between each MRI versus MATH data than between each MRI versus HSM data. All ICC values were >0.75 indicating that all four methods were reliable for measuring BD and that the mathematical algorithm and the two complimentary non-invasive MRI protocols could objectively and reliably estimate different types of breast tissues.

  7. Image feature analysis for classification of microcalcifications in digital mammography: neural networks and genetic algorithms

    NASA Astrophysics Data System (ADS)

    Wu, Chris Y.; Tsujii, Osamu; Freedman, Matthew T.; Mun, Seong K.

    1997-04-01

    We have developed an image feature-based algorithm to classify microcalcifications associated with benign and malignant processes in digital mammograms for the diagnosis of breast cancer. The feature-based algorithm is an alternative approach to image based method for classification of microcalcifications in digital mammograms. Microcalcifications can be characterized by a number of quantitative variables describing the underling key features of a suspicious region such as the size, shape, and number of microcalcifications in a cluster. These features are calculated by an automated extraction scheme for each of the selected regions. The features are then used as input to a backpropagation neural network to make a decision regarding the probability of malignancy of a selected region. The initial selection of image features set is a rough estimation that may include redundant and non-discriminant features. A genetic algorithm is employed to select an optimal image feature set from the initial feature set and select an optimized structure of the neural network for the optimal input features. The performance of neural network is compared with that of radiologists in classifying the clusters of microcalcifications. Two set of mammogram cases are used in this study. The first set is from the digital mammography database from the Mammographic Image Analysis Society (MIAS). The second set is from cases collected at Georgetown University Medical Center (GUMC). The diagnostic truth of the cases have been verified by biopsy. The performance of the neural network system is evaluated by ROC analysis. The system of neural network and genetic algorithms improves performance of our previous TRBF neural network. The neural network system was able to classify benign and malignant microcalcifications at a level favorably compared to experienced radiologists. The use of the neural network system can be used to help radiologists reducing the number biopsies in clinical applications. Genetic algorithms are an effective tool to select optimal input features and structure of a backpropagation neural network. The neural network, combined with genetic algorithms, is able to effectively classify benign and malignant microcalcifications. The results of the neural network system can be used to help reducing the number of benign biopsies.

  8. A comparison of lesion detection accuracy using digital mammography and flat-panel CT breast imaging (Honorable Mention Poster Award)

    NASA Astrophysics Data System (ADS)

    Gong, Xing; Vedula, Aruna A.; Thacker, Samta; Glick, Stephen J.

    2005-04-01

    Although conventional mammography is currently the best modality to detect early breast cancer, it is limited in that the recorded image represents the superposition of a 3D object onto a 2D plane. As an alternative, cone-beam CT breast imaging with a CsI based flat-panel imager (CTBI) has been proposed with the ability to provide 3D visualization of breast tissue. To investigate possible improvements in lesion detection accuracy using CTBI over digital mammography (DM), a computer simulation study was conducted using simulated lesions embedded into a structured 3D breast model. The computer simulation realistically modeled x-ray transport through a breast model, as well as the signal and noise propagation through the flat-panel imager. Polyenergetic x-ray spectra of W/Al 50 kVp for CTBI and Mo/Mo 28 kVp for DM were modeled. For the CTBI simulation, the intensity of the x-ray spectra for each projection view was determined so as to provide a total mean glandular dose (MGD) of 4 mGy, which is approximately equivalent to that given in a conventional two-view screening mammography study. Since only one DM view was investigated here, the intensity of the DM x-ray spectra was defined to give 2 mGy MGD. Irregular lesions were simulated by using a stochastic growth algorithm providing lesions with an effective diameter of 5 mm. Breast tissue was simulated by generating an ensemble of backgrounds with a power law spectrum. To evaluate lesion detection accuracy, a receiver operating characteristic (ROC) study was performed with 4 observers reading an ensemble of images for each case. The average area under the ROC curves (Az) was 0.94 for CTBI, and 0.81 for DM. Results indicate that a 5 mm lesion embedded in a structured breast phantom can be detected by CT breast imaging with statistically significant higher confidence than with digital mammography.

  9. Imaging lobular breast carcinoma: comparison of synchrotron radiation DEI-CT technique with clinical CT, mammography and histology

    NASA Astrophysics Data System (ADS)

    Fiedler, S.; Bravin, A.; Keyriläinen, J.; Fernández, M.; Suortti, P.; Thomlinson, W.; Tenhunen, M.; Virkkunen, P.; Karjalainen-Lindsberg, M.-L.

    2004-01-01

    Different modalities for imaging cancer-bearing breast tissue samples are described and compared. The images include clinical mammograms and computed tomography (CT) images, CT images with partly coherent synchrotron radiation (SR), and CT and radiography images taken with SR using the diffraction enhanced imaging (DEI) method. The images are evaluated by a radiologist and compared with histopathological examination of the samples. Two cases of lobular carcinoma are studied in detail. The indications of cancer are very weak or invisible in the conventional images, but the morphological changes due to invasion of cancer become pronounced in the images taken by the DEI method. The strands penetrating adipose tissue are seen clearly in the DEI-CT images, and the histopathology confirms that some strands contain the so-called 'Indian file' formations of cancer cells. The radiation dose is carefully measured for each of the imaging modalities. The mean glandular dose (MGD) for 50% glandular breast tissue is about 1 mGy in conventional mammography and less than 0.25 mGy in projection DEI, while in the clinical CT imaging the MGD is very high, about 45 mGy. The entrance dose of 95 mGy in DEI-CT imaging gives rise to an MGD of 40 mGy, but the dose may be reduced by an order of magnitude, because the contrast is very large in most images.

  10. Imaging lobular breast carcinoma: comparison of synchrotron radiation DEI-CT technique with clinical CT, mammography and histology.

    PubMed

    Fiedler, S; Bravin, A; Keyriläinen, J; Fernández, M; Suortti, P; Thomlinson, W; Tenhunen, M; Virkkunen, P; Karjalainen-Lindsberg, M

    2004-01-21

    Different modalities for imaging cancer-bearing breast tissue samples are described and compared. The images include clinical mammograms and computed tomography (CT) images, CT images with partly coherent synchrotron radiation (SR), and CT and radiography images taken with SR using the diffraction enhanced imaging (DEI) method. The images are evaluated by a radiologist and compared with histopathological examination of the samples. Two cases of lobular carcinoma are studied in detail. The indications of cancer are very weak or invisible in the conventional images, but the morphological changes due to invasion of cancer become pronounced in the images taken by the DEI method. The strands penetrating adipose tissue are seen clearly in the DEI-CT images, and the histopathology confirms that some strands contain the so-called 'Indian file' formations of cancer cells. The radiation dose is carefully measured for each of the imaging modalities. The mean glandular dose (MGD) for 50% glandular breast tissue is about 1 mGy in conventional mammography and less than 0.25 mGy in projection DEI, while in the clinical CT imaging the MGD is very high, about 45 mGy. The entrance dose of 95 mGy in DEI-CT imaging gives rise to an MGD of 40 mGy, but the dose may be reduced by an order of magnitude, because the contrast is very large in most images. PMID:15083665

  11. Projection-based energy weighting on photon-counting X-ray images in digital subtraction mammography: a feasibility study

    NASA Astrophysics Data System (ADS)

    Choi, Sung-Hoon; Lee, Seung-Wan; Choi, Yu-Na; Lee, Young-Jin; Kim, Hee-Joung

    2014-03-01

    In digital subtraction mammography where subtracts the one image (with contrast medium) from the other (anatomical background) for observing the tumor structure, tumors which include more blood vessels than normal tissue could be distinguished through the enhancement of contrast-to-noise ratio (CNR). In order to improve CNR, we adopted projection-based energy weighting for iodine solutions with four different concentrations embedded in a breast phantom (50% adipose and 50% glandular tissues). In this study, a Monte Carlo simulation was used to simulate a 40 mm thickness breast phantom, which has 15 and 30 mg/cm3 iodine solutions with two different thicknesses, and an energy resolving photon-counting system. The input energy spectrum was simulated in a range of 20 to 45 keV in order to reject electronic noise and include k-edge energy of iodine (33.2 keV). The results showed that the projection-based energy weighting improved the CNR by factors of 1.05-1.86 compared to the conventional integrating images. Consequently, the CNR of images from the digital subtraction mammography could be improved by the projection-based energy weighting with photon-counting detectors.

  12. Visualisation of calcifications and thin collagen strands in human breast tumour specimens by the diffraction-enhanced imaging technique: a comparison with conventional mammography and histology.

    PubMed

    Keyriläinen, Jani; Fernández, Manuel; Fiedler, Stefan; Bravin, Alberto; Karjalainen-Lindsberg, Marja-Liisa; Virkkunen, Pekka; Elo, Eva-Maria; Tenhunen, Mikko; Suortti, Pekka; Thomlinson, William

    2005-02-01

    Six excised human breast tissue specimens carrying benign and malignant tumours were examined with the diffraction-enhanced imaging technique. Diffraction-enhanced images were compared with diagnostic screen-film mammograms and the correlation with histological information of the specimens was established. The enhanced visibility of calcifications, some of which were smaller than 0.15 mm in diameter, is reported in detail. Fine details of the structures such as strands of collagen and contours between glandular and adipose tissue, which are barely visible at the contrast detection limit in the conventional absorption-based mammograms, are clearly visible in the diffraction-enhanced images. Microscopic study of the stained histopathological sections unequivocally confirms the correlation of the radiographic findings with the morphologic changes in specimens. An increased soft tissue contrast and a combination of information obtained with disparate diffraction-enhanced images provide better visibility of mammographically indistinguishable features. This kind of additional structural information of the breast tissue is required to improve assessment accuracy and earlier detection of the breast lesions. These advances in image quality make the method a very promising candidate for mammography. PMID:15664286

  13. A laser-driven undulator x-ray source: simulation of image formation and dose deposition in mammography

    NASA Astrophysics Data System (ADS)

    Müller, Bernhard; Schlattl, Helmut; Grüner, Florian; Hoeschen, Christoph

    2011-03-01

    Since overcoming some of the inherent limitations of x-ray tubes becomes increasingly harder, it is important to consider new ways of x-ray generation and to study their applications in the field of medical imaging. In the present work we investigate a novel table-top-sized x-ray source, developed in a joint project within the Cluster of Excellence "Munich Center for Advanced Photonics". It uses laser-accelerated electrons emitting x-ray radiation in a short period undulator. This source has the potential to deliver tunable x-rays with a very narrow spectral bandwidth. The main purpose of this contribution is to investigate the performance of this source in the field of mammography and to compare it to that of conventional x-ray tubes. We simulated the whole imaging process from the electron beam dynamics through the generation of the synchrotron radiation in the undulator up to the x-ray-matter interaction and detection in the mammographic setting. A Monte Carlo simulation of the absorption and scattering processes based on the Geant4 software toolkit has been developed that uses a high-resolution voxel phantom of the female breast for the accurate simulation of mammography. We present simulated mammograms generated by using quasi-monochromatic undulator radiation and by using the polychromatic spectrum of a conventional x-ray tube.

  14. Linear image reconstruction for a diffuse optical mammography system in a noncompressed geometry using scattering fluid

    SciTech Connect

    Nielsen, Tim; Brendel, Bernhard; Ziegler, Ronny; Beek, Michiel van; Uhlemann, Falk; Bontus, Claas; Koehler, Thomas

    2009-04-01

    Diffuse optical tomography (DOT) is a potential new imaging modality to detect or monitor breast lesions. Recently, Philips developed a new DOT system capable of transmission and fluorescence imaging, where the investigated breast is hanging freely into the measurement cup containing scattering fluid. We present a fast and robust image reconstruction algorithm that is used for the transmission measurements. The algorithm is based on the Rytov approximation. We show that this algorithm can be used over a wide range of tissue optical properties if the reconstruction is adapted to each patient. We use estimates of the breast shape and average tissue optical properties to initialize the reconstruction, which improves the image quality significantly. We demonstrate the capability of the measurement system and reconstruction to image breast lesions by clinical examples.

  15. Study of quality perception in medical images based on comparison of contrast enhancement techniques in mammographic images

    NASA Astrophysics Data System (ADS)

    Matheus, B.; Verçosa, L. B.; Barufaldi, B.; Schiabel, H.

    2014-03-01

    With the absolute prevalence of digital images in mammography several new tools became available for radiologist; such as CAD schemes, digital zoom and contrast alteration. This work focuses in contrast variation and how the radiologist reacts to these changes when asked to evaluated image quality. Three contrast enhancing techniques were used in this study: conventional equalization, CCB Correction [1] - a digitization correction - and value subtraction. A set of 100 images was used in tests from some available online mammographic databases. The tests consisted of the presentation of all four versions of an image (original plus the three contrast enhanced images) to the specialist, requested to rank each one from the best up to worst quality for diagnosis. Analysis of results has demonstrated that CCB Correction [1] produced better images in almost all cases. Equalization, which mathematically produces a better contrast, was considered the worst for mammography image quality enhancement in the majority of cases (69.7%). The value subtraction procedure produced images considered better than the original in 84% of cases. Tests indicate that, for the radiologist's perception, it seems more important to guaranty full visualization of nuances than a high contrast image. Another result observed is that the "ideal" scanner curve does not yield the best result for a mammographic image. The important contrast range is the middle of the histogram, where nodules and masses need to be seen and clearly distinguished.

  16. A dual-energy subtraction technique for microcalcification imaging in digital mammography--a signal-to-noise analysis.

    PubMed

    Lemacks, Michael R; Kappadath, S Cheenu; Shaw, Chris C; Liu, Xinming; Whitman, Gary J

    2002-08-01

    Breast cancer may manifest as microcalcifications (microCs) in x-ray mammography. However, the detection and visualization of microCs are often obscured by the overlapping tissue structures. The dual-energy subtraction imaging technique offers an alternative approach for imaging and visualizing microCs. With this technique, separate high- and low-energy images are acquired and their differences are used to "cancel" out the background tissue structures. However, the subtraction process could increase the statistical noise level relative to the calcification contrast. Therefore, a key issue with the dual-energy subtraction imaging technique is to weigh the benefit of removing the cluttered background tissue structure over the drawback of reduced signal-to-noise ratio in the subtracted microC images. In this report, a theoretical framework for calculating the (quantum) noise in the subtraction images is developed and the numerical computations are described. We estimate the noise levels in the dual-energy subtraction signals under various imaging conditions, including the x-ray spectra, microC size, tissue composition, and breast thickness. The selection of imaging parameters is optimized to evaluate the feasibility of using a dual-energy subtraction technique for the improved detection and visualization of microCs. We present the results and discuss its dependence on imaging parameters. PMID:12201421

  17. Evolution of mammographic image quality in the state of Rio de Janeiro*

    PubMed Central

    Villar, Vanessa Cristina Felippe Lopes; Seta, Marismary Horsth De; de Andrade, Carla Lourenço Tavares; Delamarque, Elizabete Vianna; de Azevedo, Ana Cecília Pedrosa

    2015-01-01

    Objective To evaluate the evolution of mammographic image quality in the state of Rio de Janeiro on the basis of parameters measured and analyzed during health surveillance inspections in the period from 2006 to 2011. Materials and Methods Descriptive study analyzing parameters connected with imaging quality of 52 mammography apparatuses inspected at least twice with a one-year interval. Results Amongst the 16 analyzed parameters, 7 presented more than 70% of conformity, namely: compression paddle pressure intensity (85.1%), films development (72.7%), film response (72.7%), low contrast fine detail (92.2%), tumor mass visualization (76.5%), absence of image artifacts (94.1%), mammography-specific developers availability (88.2%). On the other hand, relevant parameters were below 50% conformity, namely: monthly image quality control testing (28.8%) and high contrast details with respect to microcalcifications visualization (47.1%). Conclusion The analysis revealed critical situations in terms of compliance with the health surveillance standards. Priority should be given to those mammography apparatuses that remained non-compliant at the second inspection performed within the one-year interval. PMID:25987749

  18. [Perspectives of the digital mammography platform].

    PubMed

    Gruber, R; Riedl, C C; Reisegger, M; Pinker, K; Sturm, E; Semturs, F; Helbich, T H

    2010-11-01

    In Europe one out of every nine women suffers from breast cancer during her lifetime. Since the introduction of mammography screening programs more breast cancers are being diagnosed when they are still small and early stage cancers with a favourable prognosis. The introduction of digital mammography systems has led to a continuous reduction of breast cancer mortality especially in specific patient subgroups. Furthermore, the digital mammography platform enables the development of new, innovative breast imaging methods to increase sensitivity and decrease breast cancer mortality. This digital mammography platform includes digital breast tomosynthesis, digital contrast medium mammography and digital contrast medium breast tomosynthesis as well as fused data sets from digital mammography with ultrasound or MRI. The following article summarizes these new applications, describes the strengths of the digital platform and illustrates the potential advantages of an improved breast cancer diagnosis by digital mammography. PMID:20945148

  19. Dose sensitivity of three phantoms used for quality assurance in digital mammography.

    PubMed

    Figl, M; Semturs, F; Kaar, M; Hoffmann, R; Kaldarar, H; Homolka, P; Mostbeck, G; Scholz, B; Hummel, J

    2013-01-21

    Technical quality assurance (QA) is one of the key issues in breast cancer screening protocols. For this QA task, three different methods are commonly used to assess image quality. The European protocol suggests a contrast-detail phantom (e.g. the CDMAM phantom), while in North America the American College of Radiology (ACR) accreditation phantom is proposed. Alternatively, phantoms based on image quality parameters from applied system theory such as the noise-equivalent number of quanta (NEQ) are applied (e.g. the PAS 1054 phantom). The aim of this paper was to correlate the changes in the output of the three evaluation methods (CDMAM, ACR and NEQ) with changes in dose. We varied the time-current product within a range of clinically used values (40-140 mAs, corresponding to 3.5-12.4 mGy entrance dose and detector dose of 32-110 ?Gy). For the ACR phantom, the examined parameter was the number of detected objects. With the CDMAM phantom we chose the diameters 0.10, 0.13, 0.20, 0.31 and 0.5 mm and recorded the threshold thicknesses. With respect to the third method, we evaluated the NEQ at typical spatial frequencies to calculate the relative changes in NEQ. Plotting NEQ versus dose increment shows a linear relationship and can be described by a linear function (with R > 0.99). Every manually selectable current- time product increment can be detected. With the ACR phantom, the number of detected objects increases only in the lower dose range and reaches saturation at about 9 mGy entrance dose (80 ?Gy detector dose). The CDMAM can detect a 50% increase in dose over the examined dose range with all five diameters, although the increases of threshold thickness are not monotonous. We conclude that an NEQ-based method has the potential to replace the established detail phantom methods to detect dose changes in the course of QA. PMID:23257608

  20. Dose sensitivity of three phantoms used for quality assurance in digital mammography

    NASA Astrophysics Data System (ADS)

    Figl, M.; Semturs, F.; Kaar, M.; Hoffmann, R.; Kaldarar, H.; Homolka, P.; Mostbeck, G.; Scholz, B.; Hummel, J.

    2013-01-01

    Technical quality assurance (QA) is one of the key issues in breast cancer screening protocols. For this QA task, three different methods are commonly used to assess image quality. The European protocol suggests a contrast-detail phantom (e.g. the CDMAM phantom), while in North America the American College of Radiology (ACR) accreditation phantom is proposed. Alternatively, phantoms based on image quality parameters from applied system theory such as the noise-equivalent number of quanta (NEQ) are applied (e.g. the PAS 1054 phantom). The aim of this paper was to correlate the changes in the output of the three evaluation methods (CDMAM, ACR and NEQ) with changes in dose. We varied the time-current product within a range of clinically used values (40-140 mAs, corresponding to 3.5-12.4 mGy entrance dose and detector dose of 32-110 μGy). For the ACR phantom, the examined parameter was the number of detected objects. With the CDMAM phantom we chose the diameters 0.10, 0.13, 0.20, 0.31 and 0.5 mm and recorded the threshold thicknesses. With respect to the third method, we evaluated the NEQ at typical spatial frequencies to calculate the relative changes in NEQ. Plotting NEQ versus dose increment shows a linear relationship and can be described by a linear function (with R > 0.99). Every manually selectable current- time product increment can be detected. With the ACR phantom, the number of detected objects increases only in the lower dose range and reaches saturation at about 9 mGy entrance dose (80 μGy detector dose). The CDMAM can detect a 50% increase in dose over the examined dose range with all five diameters, although the increases of threshold thickness are not monotonous. We conclude that an NEQ-based method has the potential to replace the established detail phantom methods to detect dose changes in the course of QA.

  1. Comparison of in vitro breast cancer visibility in analyser-based computed tomography with histopathology, mammography, computed tomography and magnetic resonance imaging.

    PubMed

    Keyriläinen, Jani; Fernández, Manuel; Bravin, Alberto; Karjalainen-Lindsberg, Marja Liisa; Leidenius, Marjut; von Smitten, Karl; Tenhunen, Mikko; Kangasmäki, Aki; Sipilä, Petri; Nemoz, Christian; Virkkunen, Pekka; Suortti, Pekka

    2011-09-01

    High-resolution analyser-based X-ray imaging computed tomography (HR ABI-CT) findings on in vitro human breast cancer are compared with histopathology, mammography, computed tomography (CT) and magnetic resonance imaging. The HR ABI-CT images provided significantly better low-contrast visibility compared with the standard radiological images. Fine cancer structures indistinguishable and superimposed in mammograms were seen, and could be matched with the histopathological results. The mean glandular dose was less than 1 mGy in mammography and 12-13 mGy in CT and ABI-CT. The excellent visibility of in vitro breast cancer suggests that HR ABI-CT may have a valuable role in the future as an adjunct or even alternative to current breast diagnostics, when radiation dose is further decreased, and compact synchrotron radiation sources become available. PMID:21862846

  2. The Values of Combined and Sub-Stratified Imaging Scores with Ultrasonography and Mammography in Breast Cancer Subtypes

    PubMed Central

    Chang, Tsun-Hou; Hsu, Hsian-He; Chou, Yu-Ching; Yu, Jyh-Cherng; Hsu, Giu-Cheng; Huang, Guo-Shu; Liao, Guo-Shiou

    2015-01-01

    Background and Objectives The Breast Imaging Reporting and Data System (BI-RADS) of Mammography (MG) and Ultrasonography (US) were equivalent to the “5-point score” and applied for combined and sub-stratified imaging assessments. This study evaluated the value of combined and sub-stratified imaging assessments with MG and US over breast cancer subtypes (BCS). Materials and Methods Medical records of 5,037 cases having imaging-guided core biopsy, performed from 2009 to 2012, were retrospectively reviewed. This study selected 1,995 cases (1,457 benign and 538 invasive cancer) having both MG and US before biopsy. These cases were categorized with the “5-point score” for their MG and US, and applied for combined and sub-stratified imaging assessments. Invasive cancers were classified on the basis of BCS, and correlated with combined and sub-stratified imaging assessments. Results These selected cases were evaluated by the “5-point score.” MG, US, and combined and sub-stratified imaging assessments all revealed statistically significant (P < 0.001) incidence of malignancy. The sensitivity was increased in the combined imaging score (99.8%), and the specificity was increased in the sub-stratified combined score (75.4%). In the sub-stratified combined imaging assessment, all BCS can be classified with higher scores (abnormality hierarchy), and luminal B subtype showed the most salient result (hierarchy: higher, 95%; lower, 5%). Conclusions Combined and sub-stratified imaging assessments can increase sensitivity and specificity of breast cancer diagnosis, respectively, and Luminal B subtype shows the best identification by sub-stratified combined imaging scoring. PMID:26689198

  3. Image quality concepts for PACS

    NASA Astrophysics Data System (ADS)

    Glicksman, Robert A.; Prior, Fred W.

    1994-05-01

    The complexity of modern Picture Archiving and Communication Systems (PACS) confronts the potential user with a bewildering array of specifications, of which those effecting image quality are of primary importance. This paper reviews some of the basic concepts of PACS image acquisition and display and the relationship of the specifications for each to image quality. The key parameters of images quality for image digitization devices are spatial resolution, dynamic range, pixel accuracy and signal to noise ratio. On the output side critical factors are brightness, spatial and contrast resolution, stability, and uniformity. The concept of frame buffer depth vs. display depth for CRT monitors is reviewed, and the correspondence of CRT images to film based images is discussed. The principles of sampling theory and the Nyquist limit are also discussed. Through an understanding of the concepts presented in this paper, the PACS user (or potential user) will be in a better position to evaluate PACS for his/her clinical application.

  4. 75 FR 11542 - Agency Information Collection Activities; Proposed Collection; Comment Request; Mammography...

    Federal Register 2010, 2011, 2012, 2013, 2014

    2010-03-11

    ...; and standards for mammography equipment, personnel, and practices, including quality assurance. The..., quality assurance and quality control standards, and have a medical reporting and recordkeeping program, a... Collection; Comment Request; Mammography Quality Standards Act Requirements AGENCY: Food and...

  5. Image enhancement, image quality, and noise

    NASA Astrophysics Data System (ADS)

    Rahman, Zia-ur; Jobson, Daniel J.; Woodell, Glenn A.; Hines, Glenn D.

    2005-08-01

    The Multiscale Retinex With Color Restoration (MSRCR) is a non-linear image enhancement algorithm that provides simultaneous dynamic range compression, color constancy and rendition. The overall impact is to brighten up areas of poor contrast/lightness but not at the expense of saturating areas of good contrast/brightness. The downside is that with the poor signal-to-noise ratio that most image acquisition devices have in dark regions, noise can also be greatly enhanced thus affecting overall image quality. In this paper, we will discuss the impact of the MSRCR on the overall quality of an enhanced image as a function of the strength of shadows in an image, and as a function of the root-mean-square (RMS) signal-to-noise (SNR) ratio of the image.

  6. Image Enhancement, Image Quality, and Noise

    NASA Technical Reports Server (NTRS)

    Rahman, Zia-ur; Jobson, Daniel J.; Woodell, Glenn A.; Hines, Glenn D.

    2005-01-01

    The Multiscale Retinex With Color Restoration (MSRCR) is a non-linear image enhancement algorithm that provides simultaneous dynamic range compression, color constancy and rendition. The overall impact is to brighten up areas of poor contrast/lightness but not at the expense of saturating areas of good contrast/brightness. The downside is that with the poor signal-to-noise ratio that most image acquisition devices have in dark regions, noise can also be greatly enhanced thus affecting overall image quality. In this paper, we will discuss the impact of the MSRCR on the overall quality of an enhanced image as a function of the strength of shadows in an image, and as a function of the root-mean-square (RMS) signal-to-noise (SNR) ratio of the image.

  7. Calibrated breast density methods for full field digital mammography: A system for serial quality control and inter-system generalization

    PubMed Central

    Lu, B.; Smallwood, A. M.; Sellers, T. A.; Drukteinis, J. S.; Heine, J. J.

    2015-01-01

    Purpose: The authors are developing a system for calibrated breast density measurements using full field digital mammography (FFDM). Breast tissue equivalent (BTE) phantom images are used to establish baseline (BL) calibration curves at time zero. For a given FFDM unit, the full BL dataset is comprised of approximately 160 phantom images, acquired prior to calibrating prospective patient mammograms. BL curves are monitored serially to ensure they produce accurate calibration and require updating when calibration accuracy degrades beyond an acceptable tolerance, rather than acquiring full BL datasets repeatedly. BL updating is a special case of generalizing calibration datasets across FFDM units, referred to as cross-calibration. Serial monitoring, BL updating, and cross-calibration techniques were developed and evaluated. Methods: BL curves were established for three Hologic Selenia FFDM units at time zero. In addition, one set of serial phantom images, comprised of equal proportions of adipose and fibroglandular BTE materials (50/50 compositions) of a fixed height, was acquired biweekly and monitored with the cumulative sum (Cusum) technique. These 50/50 composition images were used to update the BL curves when the calibration accuracy degraded beyond a preset tolerance of ±4 standardized units. A second set of serial images, comprised of a wide-range of BTE compositions, was acquired biweekly to evaluate serial monitoring, BL updating, and cross-calibration techniques. Results: Calibration accuracy can degrade serially and is a function of acquisition technique and phantom height. The authors demonstrated that all heights could be monitored simultaneously while acquiring images of a 50/50 phantom with a fixed height for each acquisition technique biweekly, translating into approximately 16 image acquisitions biweekly per FFDM unit. The same serial images are sufficient for serial monitoring, BL updating, and cross-calibration. Serial calibration accuracy was maintained within ±4 standardized unit variation from the ideal when applying BL updating. BL updating is a special case of cross-calibration; the BL dataset of unit 1 can be converted to the BL dataset for another similar unit (i.e., unit 2) at any given time point using the 16 serial monitoring 50/50 phantom images of unit 2 (or vice versa) acquired near this time point while maintaining the ±4 standardized unit tolerance. Conclusions: A methodology for monitoring and maintaining serial calibration accuracy for breast density measurements was evaluated. Calibration datasets for a given unit can be translated forward in time with minimal phantom imaging effort. Similarly, cross-calibration is a method for generalizing calibration datasets across similar units without additional phantom imaging. This methodology will require further evaluation with mammograms for complete validation. PMID:25652480

  8. Retinal Image Quality During Accommodation

    PubMed Central

    López-Gil, N.; Martin, J.; Liu, T.; Bradley, A.; Díaz-Muñoz, D.; Thibos, L.

    2013-01-01

    Purpose We asked if retinal image quality is maximum during accommodation, or sub-optimal due to accommodative error, when subjects perform an acuity task. Methods Subjects viewed a monochromatic (552nm), high-contrast letter target placed at various viewing distances. Wavefront aberrations of the accommodating eye were measured near the endpoint of an acuity staircase paradigm. Refractive state, defined as the optimum target vergence for maximising retinal image quality, was computed by through-focus wavefront analysis to find the power of the virtual correcting lens that maximizes visual Strehl ratio. Results Despite changes in ocular aberrations and pupil size during binocular viewing, retinal image quality and visual acuity typically remain high for all target vergences. When accommodative errors lead to sub-optimal retinal image quality, acuity and measured image quality both decline. However, the effect of accommodation errors of on visual acuity are mitigated by pupillary constriction associated with accommodation and binocular convergence and also to binocular summation of dissimilar retinal image blur. Under monocular viewing conditions some subjects displayed significant accommodative lag that reduced visual performance, an effect that was exacerbated by pharmacological dilation of the pupil. Conclusions Spurious measurement of accommodative error can be avoided when the image quality metric used to determine refractive state is compatible with the focusing criteria used by the visual system to control accommodation. Real focusing errors of the accommodating eye do not necessarily produce a reliably measurable loss of image quality or clinically significant loss of visual performance, probably because of increased depth-of-focus due to pupil constriction. When retinal image quality is close to maximum achievable (given the eye’s higher-order aberrations), acuity is also near maximum. A combination of accommodative lag, reduced image quality, and reduced visual function may be a useful sign for diagnosing functionally-significant accommodative errors indicating the need for therapeutic intervention. PMID:23786386

  9. Screening for breast cancer with mammography

    SciTech Connect

    Sickles, E.A. )

    1991-10-01

    Mammography is generally accepted as a useful problem-solving clinical tool in characterizing known breast lesions, so that appropriate and timely treatment can be given. However, it remains grossly underutilized at what it does best: screening. The major strengths of mammography are (a) its ability to detect breast cancer at a smaller, potentially more curable stage than any other examination, and (b) its proved efficacy in reducing breast cancer mortality in asymptomatic women aged 40-74. If, as has recently been estimated, screening with mammography and physical examination can be expected to lower breast cancer deaths by 40%-50% among those actually examined (13), then the lives of almost 20,000 U.S. women might be saved each year if screening were to become very widely used. The challenges of the next decade are clear, to mount much more effective campaigns to educate physicians and lay women about the life-saving benefits of breast cancer screening, to devise increasingly effective and lower cost screening strategies, to further improve the current high quality of mammographic imaging despite its increasing proliferation, and to train large numbers of breast imaging specialists to guarantee that the growing case load of screening and problem-solving mammograms is interpreted with a very high level of skill.

  10. Quality assessment for hyperspectral imaging

    NASA Astrophysics Data System (ADS)

    Chen, Yuheng; Chen, Xinhua; Zhou, Jiankang; Shen, Weimin

    2014-11-01

    Image quality assessment is an essential value judgement approach for many applications. Multi & hyper spectral imaging has more judging essentials than grey scale or RGB imaging and its image quality assessment job has to cover up all-around evaluating factors. This paper presents an integrating spectral imaging quality assessment project, in which spectral-based, radiometric-based and spatial-based statistical behavior for three hyperspectral imagers are jointly executed. Spectral response function is worked out based on discrete illumination images and its spectral performance is deduced according to its FWHM and spectral excursion value. Radiometric response ability of different spectral channel under both on-ground and airborne imaging condition is judged by SNR computing based upon local RMS extraction and statistics method. Spatial response evaluation of the spectral imaging instrument is worked out by MTF computing with slanted edge analysis method. Reported pioneering systemic work in hyperspectral imaging quality assessment is carried out with the help of several domestic dominating work units, which not only has significance in the development of on-ground and in-orbit instrument performance evaluation technique but also takes on reference value for index demonstration and design optimization for instrument development.

  11. Breast Cancer: Comparative Effectiveness of Positron Emission Mammography and MR Imaging in Presurgical Planning for the Ipsilateral Breast1

    PubMed Central

    Madsen, Kathleen S.; Schilling, Kathy; Tartar, Marie; Pisano, Etta D.; Larsen, Linda Hovanessian; Narayanan, Deepa; Ozonoff, Al; Miller, Joel P.; Kalinyak, Judith E.

    2011-01-01

    Purpose: To determine the performance of positron emission mammography (PEM), as compared with magnetic resonance (MR) imaging, including the effect on surgical management, in ipsilateral breasts with cancer. Materials and Methods: Four hundred seventy-two women with newly diagnosed breast cancer who were offered breast-conserving surgery consented from September 2006 to November 2008 to participate in a multicenter institutional review board–approved, HIPAA-compliant protocol. Participants underwent contrast material–enhanced MR imaging and fluorine 18 fluorodeoxyglucose PEM in randomized order; resultant images were interpreted independently. Added biopsies and changes in surgical procedure for the ipsilateral breast were correlated with histopathologic findings. Performance characteristics were compared by using the McNemar test and generalized estimating equations. Results: Three hundred eighty-eight women (median age, 58 years; age range, 26–93 years; median estimated tumor size, 1.5 cm) completed the study. Additional cancers were found in 82 (21%) women (82 ipsilateral breasts; median tumor size, 0.7 cm). Twenty-eight (34%) of the 82 breasts were identified with both PEM and MR imaging; 21 (26%) breasts, with MR imaging only; 14 (17%) breasts, with PEM only; and seven (8.5%) breasts, with mammography and ultrasonography. Twelve (15%) cases of additional cancer were missed at all imaging examinations. Integration of PEM and MR imaging increased cancer detection—to 61 (74%) of 82 breasts versus 49 (60%) of 82 breasts identified with MR imaging alone (P < .001). Of 306 breasts without additional cancer, 279 (91.2%) were correctly assessed with PEM compared with 264 (86.3%) that were correctly assessed with MR imaging (P = .03). The positive predictive value of biopsy prompted by PEM findings (47 [66%] of 71 cases) was higher than that of biopsy prompted by MR findings (61 [53%] of 116 cases) (P = .016). Of 116 additional cancers, 61 (53%) were depicted by MR imaging and 47 (41%) were depicted by PEM (P = .043). Fifty-six (14%) of the 388 women required mastectomy: 40 (71%) of these women were identified with MR imaging, and 20 (36%) were identified with PEM (P < .001). Eleven (2.8%) women underwent unnecessary mastectomy, which was prompted by only MR findings in five women, by only PEM findings in one, and by PEM and MR findings in five. Thirty-three (8.5%) women required wider excision: 24 (73%) of these women were identified with MR imaging, and 22 (67%) were identified with PEM. Conclusion: PEM and MR imaging had comparable breast-level sensitivity, although MR imaging had greater lesion-level sensitivity and more accurately depicted the need for mastectomy. PEM had greater specificity at the breast and lesion levels. Eighty-nine (23%) participants required more extensive surgery: 61 (69%) of these women were identified with MR imaging, and 41 (46%) were identified with PEM (P = .003). Fourteen (3.6%) women had tumors seen only at PEM. © RSNA, 2010 Supplemental material: http://radiology.rsna.org/lookup/suppl/doi:10.1148/radiol.10100454/-/DC1 PMID:21076089

  12. Design and analysis of a focused mammography imaging system with suppressed background radiation

    NASA Astrophysics Data System (ADS)

    Nahm, Kie B.; Mun, Seong K.

    1996-04-01

    A design concept of a mammography system is presented that is expected to reduce the background radiation without the penalty of the increased dosage or time. The monochromatic radiation from Mo target is Bragg reflected from a crystal mirror fabricated as per the Johansson geometry. Radiation then comes to a focus on the opposite side of the source point, and a pinhole located at this focal point will filter out the scattering from the bulk of the breast tissue. The focal spot size should be identical to the target spot size under an ideal circumstance. LiF crystal was assumed for the study, but any crystal with appropriate lattice constant and reflectivity can be adopted, depending on the difficulty of fabrication. Expected technical difficulties are presented and discussed.

  13. Recent advances in screen-film mammography

    SciTech Connect

    Haus, A.G.

    1987-09-01

    Today there are many dedicated mammographic x-ray units available that are capable of providing high-quality screen-film mammograms. Likewise, screen-film combinations designed for mammography are capable of providing images with appropriate contrast, resolution, and noise levels. Proper film processing is most important in order to obtain the appropriate film speed and contrast. A higher-speed screen-film combination designed for mammography can provide mammograms with significantly lower radiation dose, especially for grid and magnification techniques. Designing x-ray units and techniques as well as screen-film combinations with the singular goal of reducing radiation dose will always involve compromises and trade-offs. The key is to always consider optimizing all of the factors that affect image quality: (1) appropriate beam quality, (2) breast compression, (3) consideration of the use of grids, (4) good geometry, (5) selection of an appropriate screen-film combination, and (6) proper film processing. Optimization of all appropriate imaging factors will produce high-quality mammograms at the lowest radiation dose to the patient.52 references.

  14. Digital breast tomosynthesis: application of 2D digital mammography CAD to detection of microcalcification clusters on planar projection image

    NASA Astrophysics Data System (ADS)

    Samala, Ravi K.; Chan, Heang-Ping; Lu, Yao; Hadjiiski, Lubomir; Wei, Jun; Helvie, Mark

    2015-03-01

    Computer-aided detection (CAD) has the potential to aid radiologists in detection of microcalcification clusters (MCs). CAD for digital breast tomosynthesis (DBT) can be developed by using the reconstructed volume, the projection views or other derivatives as input. We have developed a novel method of generating a single planar projection (PPJ) image from a regularized DBT volume to emphasize the high contrast objects such as microcalcifications while removing the anatomical background and noise. In this work, we adapted a CAD system developed for digital mammography (CADDM) to the PPJ image and compared its performance with our CAD system developed for DBT volumes (CADDBT) in the same set of cases. For microcalcification detection in the PPJ image using the CADDM system, the background removal preprocessing step designed for DM was not needed. The other methods and processing steps in the CADDM system were kept without modification while the parameters were optimized with a training set. The linear discriminant analysis classifier using cluster based features was retrained to generate a discriminant score to be used as decision variable. For view-based FROC analysis, at 80% sensitivity, an FP rate of 1.95/volume and 1.54/image were achieved, respectively, for CADDBT and CADDM in an independent test set. At a threshold of 1.2 FPs per image or per DBT volume, the nonparametric analysis of the area under the FROC curve shows that the optimized CADDM for PPJ is significantly better than CADDBT. However, the performance of CADDM drops at higher sensitivity or FP rate, resulting in similar overall performance between the two CAD systems. The higher sensitivity of the CADDM in the low FP rate region and vice versa for the CADDBT indicate that a joint CAD system combining detection in the DBT volume and the PPJ image has the potential to increase the sensitivity and reduce the FP rate.

  15. Combined Optical Imaging and Mammography of the Healthy Breast: Optical Contrast Derived From Breast Structure and Compression

    PubMed Central

    Fang, Qianqian; Carp, Stefan A.; Selb, Juliette; Boverman, Greg; Zhang, Quan; Kopans, Daniel B.; Moore, Richard H.; Miller, Eric L.; Brooks, Dana H.; Boas, David A.

    2009-01-01

    In this paper, we report new progress in developing the instrument and software platform of a combined X-ray mammography/diffuse optical breast imaging system. Particularly, we focus on system validation using a series of balloon phantom experiments and the optical image analysis of 49 healthy patients. Using the finite-element method for forward modeling and a regularized Gauss-Newton method for parameter reconstruction, we recovered the inclusions inside the phantom and the hemoglobin images of the human breasts. An enhanced coupling coefficient estimation scheme was also incorporated to improve the accuracy and robustness of the reconstructions. The recovered average total hemoglobin concentration (HbT) and oxygen saturation (SO2) from 68 breast measurements are 16.2 ?m and 71%, respectively, where the HbT presents a linear trend with breast density. The low HbT value compared to literature is likely due to the associated mammographic compression. From the spatially co-registered optical/X-ray images, we can identify the chest-wall muscle, fatty tissue, and fibroglandular regions with an average HbT of 20.1±6.1 ?m for fibroglandular tissue, 15.4±5.0 ?m for adipose, and 22.2±7.3 ?m for muscle tissue. The differences between fibroglandular tissue and the corresponding adipose tissue are significant (p < 0.0001). At the same time, we recognize that the optical images are influenced, to a certain extent, by mammographical compression. The optical images from a subset of patients show composite features from both tissue structure and pressure distribution. We present mechanical simulations which further confirm this hypothesis. PMID:19116186

  16. Analysis of multilayer and single layer X-ray detectors for contrast-enhanced mammography using imaging task

    NASA Astrophysics Data System (ADS)

    Allec, Nicholas; Abbaszadeh, Shiva; Karim, Karim S.

    2011-03-01

    A multilayer (single-shot) detector has previously been proposed for contrast-enhanced mammography. The multilayer detector has the benefit of avoiding motion artifacts due to simultaneous acquisition of both high and low energy images. A single layer (dual-shot) detector has the benefit of better control over the energy separation since the incident beams can be produced and filtered separately. In this paper the performance of the multilayer detector is compared to that of a single layer detector using an ideal observer detectability index which is determined from an extended cascaded systems model and a defined imaging task. The detectors are assumed to have amorphous selenium direct conversion layers, however the same theoretical techniques used here may be applied to other types of integrating detectors. The anatomical noise caused by variation of glandularity within the breast is known to dominate the noise power spectrum at low frequencies due to its inverse power law dependence and is thus taken into account in our model to provide an accurate estimate of the detectability index. The conditions leading to the optimal detectability index, such as tube voltage, filtration, and weight factor are reported for both detector designs.

  17. The imaging performance of compact Lu2O3:Eu powdered phosphor screens: Monte Carlo simulation for applications in mammography.

    PubMed

    Liaparinos, P F; Kandarakis, I S

    2009-06-01

    In medical mammographic imaging systems, one type of detector configuration, often referred to as indirect detectors, is based on a scintillator layer (phosphor screen) that converts the x-ray radiation into optical signal. The indirect detector performance may be optimized either by improving the structural parameters of the screen or by employing new phosphor materials with improved physical characteristics (e.g., x-ray absorption efficiency, intrinsic conversion efficiency, emitted light spectrum). Lu2O3:Eu is a relatively new phosphor material that exhibits improved scintillating properties indicating a promising material for mammographic applications. In this article, a custom validated Monte Carlo program was used in order to examine the performance of compact Lu2O3:Eu powdered phosphor screens under diagnostic mammography conditions (x-ray spectra: 28 kV Mo, 0.030 mm Mo and 32 kV W, 0.050 mm Rh). Lu2O3:Eu screens of coating weight in the range between 20 and 40 mg/cm2 were examined. The Monte Carlo code was based on a model using Mie-scattering theory for the description of light propagation within the phosphor. The overall performance of Lu2O3:Eu powdered phosphor screens was investigated in terms of the (i) quantum detection efficiency, (ii) luminescence efficiency, (iii) compatibility with optical sensors, (iv) modulation transfer function, (v) the Swank factor, and (vi) zero-frequency detective quantum efficiency. Results were compared to the traditional rare-earth Gd2O2S:Tb phosphor material. The increased packing density and therefore the light extinction properties of Lu2O3:Eu phosphor were found to improve the x-ray absorption (approximately up to 21% and 16% at 40 mg/cm2 for Mo and W x-ray spectra, respectively), the spatial resolution (approximately 2.6 and 2.4 cycles/mm at 40 mg/cm2 for Mo and W x-ray spectra, respectively), as well as the zero-frequency detective quantum efficiency (approximately up to 8% and 18% at 20 mg/cm2 for Mo and W x-ray spectra, respectively) of the screens in comparison to the Gd2O2S: Tb screens. Data obtained by the simulations indicate that certain optical properties of Lu2O3:Eu make this material a promising phosphor which, under appropriate conditions, could be considered for use in x-ray mammography imagers. PMID:19610287

  18. Video and image quality

    NASA Astrophysics Data System (ADS)

    Aldridge, Jim

    1995-09-01

    This paper presents some of the results of a UK government research program into methods of improving the effectiveness of CCTV surveillance systems. The paper identifies the major components of video security systems and primary causes of unsatisfactory images. A method is outline for relating the picture detail limitations imposed by each system component on overall system performance. The paper also points out some possible difficulties arising from the use of emerging new technology.

  19. Fully Automated Quantitative Estimation of Volumetric Breast Density from Digital Breast Tomosynthesis Images: Preliminary Results and Comparison with Digital Mammography and MR Imaging.

    PubMed

    Pertuz, Said; McDonald, Elizabeth S; Weinstein, Susan P; Conant, Emily F; Kontos, Despina

    2016-04-01

    Purpose To assess a fully automated method for volumetric breast density (VBD) estimation in digital breast tomosynthesis (DBT) and to compare the findings with those of full-field digital mammography (FFDM) and magnetic resonance (MR) imaging. Materials and Methods Bilateral DBT images, FFDM images, and sagittal breast MR images were retrospectively collected from 68 women who underwent breast cancer screening from October 2011 to September 2012 with institutional review board-approved, HIPAA-compliant protocols. A fully automated computer algorithm was developed for quantitative estimation of VBD from DBT images. FFDM images were processed with U.S. Food and Drug Administration-cleared software, and the MR images were processed with a previously validated automated algorithm to obtain corresponding VBD estimates. Pearson correlation and analysis of variance with Tukey-Kramer post hoc correction were used to compare the multimodality VBD estimates. Results Estimates of VBD from DBT were significantly correlated with FFDM-based and MR imaging-based estimates with r = 0.83 (95% confidence interval [CI]: 0.74, 0.90) and r = 0.88 (95% CI: 0.82, 0.93), respectively (P < .001). The corresponding correlation between FFDM and MR imaging was r = 0.84 (95% CI: 0.76, 0.90). However, statistically significant differences after post hoc correction (α = 0.05) were found among VBD estimates from FFDM (mean ± standard deviation, 11.1% ± 7.0) relative to MR imaging (16.6% ± 11.2) and DBT (19.8% ± 16.2). Differences between VDB estimates from DBT and MR imaging were not significant (P = .26). Conclusion Fully automated VBD estimates from DBT, FFDM, and MR imaging are strongly correlated but show statistically significant differences. Therefore, absolute differences in VBD between FFDM, DBT, and MR imaging should be considered in breast cancer risk assessment. (©) RSNA, 2015 Online supplemental material is available for this article. PMID:26491909

  20. Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1)

    NASA Astrophysics Data System (ADS)

    Csete, I.; Büermann, L.; Gomola, I.

    2016-01-01

    A comparison of the air kerma standards for x-radiation qualities used in mammography was performed between the PTB and the IAEA. Two reference-class ionization chamber types Radcal RC6M and Magna A650 of the IAEA and tungsten anode based beam qualities with Mo and Al external filtrations (W+Mo, W+Al) established at both laboratories were selected for the comparison. The calibration coefficients, NK_air, were determined for the transfer chambers at the PTB in May 2015 and before and after this at the IAEA Dosimetry Laboratory. The results show good agreement, to be well within the 0.55 % standard uncertainty of the comparison. Correction factors to determine NK_air for these beam qualities based on calibration in RQR-M mammography beam qualities, established according to the IEC 61267 standard, were also calculated for the Radcal RC6M, 10X5-6M, and Magna A650 types of chambers. Main text To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

  1. Effects of exposure equalization on image signal-to-noise ratios in digital mammography: A simulation study with an anthropomorphic breast phantom

    SciTech Connect

    Liu Xinming; Lai Chaojen; Whitman, Gary J.; Geiser, William R.; Shen Youtao; Yi Ying; Shaw, Chris C.

    2011-12-15

    Purpose: The scan equalization digital mammography (SEDM) technique combines slot scanning and exposure equalization to improve low-contrast performance of digital mammography in dense tissue areas. In this study, full-field digital mammography (FFDM) images of an anthropomorphic breast phantom acquired with an anti-scatter grid at various exposure levels were superimposed to simulate SEDM images and investigate the improvement of low-contrast performance as quantified by primary signal-to-noise ratios (PSNRs). Methods: We imaged an anthropomorphic breast phantom (Gammex 169 ''Rachel,'' Gammex RMI, Middleton, WI) at various exposure levels using a FFDM system (Senographe 2000D, GE Medical Systems, Milwaukee, WI). The exposure equalization factors were computed based on a standard FFDM image acquired in the automatic exposure control (AEC) mode. The equalized image was simulated and constructed by superimposing a selected set of FFDM images acquired at 2, 1, 1/2, 1/4, 1/8, 1/16, and 1/32 times of exposure levels to the standard AEC timed technique (125 mAs) using the equalization factors computed for each region. Finally, the equalized image was renormalized regionally with the exposure equalization factors to result in an appearance similar to that with standard digital mammography. Two sets of FFDM images were acquired to allow for two identically, but independently, formed equalized images to be subtracted from each other to estimate the noise levels. Similarly, two identically but independently acquired standard FFDM images were subtracted to estimate the noise levels. Corrections were applied to remove the excess system noise accumulated during image superimposition in forming the equalized image. PSNRs over the compressed area of breast phantom were computed and used to quantitatively study the effects of exposure equalization on low-contrast performance in digital mammography. Results: We found that the highest achievable PSNR improvement factor was 1.89 for the anthropomorphic breast phantom used in this study. The overall PSNRs were measured to be 79.6 for the FFDM imaging and 107.6 for the simulated SEDM imaging on average in the compressed area of breast phantom, resulting in an average improvement of PSNR by {approx}35% with exposure equalization. We also found that the PSNRs appeared to be largely uniform with exposure equalization, and the standard deviations of PSNRs were estimated to be 10.3 and 7.9 for the FFDM imaging and the simulated SEDM imaging, respectively. The average glandular dose for SEDM was estimated to be 212.5 mrad, {approx}34% lower than that of standard AEC-timed FFDM (323.8 mrad) as a result of exposure equalization for the entire breast phantom. Conclusions: Exposure equalization was found to substantially improve image PSNRs in dense tissue regions and result in more uniform image PSNRs. This improvement may lead to better low-contrast performance in detecting and visualizing soft tissue masses and micro-calcifications in dense tissue areas for breast imaging tasks.

  2. Landsat image data quality studies

    NASA Technical Reports Server (NTRS)

    Schueler, C. F.; Salomonson, V. V.

    1985-01-01

    Preliminary results of the Landsat-4 Image Data Quality Analysis (LIDQA) program to characterize the data obtained using the Thematic Mapper (TM) instrument on board the Landsat-4 and Landsat-5 satellites are reported. TM design specifications were compared to the obtained data with respect to four criteria, including spatial resolution; geometric fidelity; information content; and image relativity to Multispectral Scanner (MSS) data. The overall performance of the TM was rated excellent despite minor instabilities and radiometric anomalies in the data. Spatial performance of the TM exceeded design specifications in terms of both image sharpness and geometric accuracy, and the image utility of the TM data was at least twice as high as MSS data. The separability of alfalfa and sugar beet fields in a TM image is demonstrated.

  3. Phase-contrast enhanced mammography: A new diagnostic tool for breast imaging

    SciTech Connect

    Wang Zhentian; Thuering, Thomas; David, Christian; Roessl, Ewald; Trippel, Mafalda; Kubik-Huch, Rahel A.; Singer, Gad; Hohl, Michael K.; Hauser, Nik; Stampanoni, Marco

    2012-07-31

    Phase contrast and scattering-based X-ray imaging can potentially revolutionize the radiological approach to breast imaging by providing additional and complementary information to conventional, absorption-based methods. We investigated native, non-fixed whole breast samples using a grating interferometer with an X-ray tube-based configuration. Our approach simultaneously recorded absorption, differential phase contrast and small-angle scattering signals. The results show that this novel technique - combined with a dedicated image fusion algorithm - has the potential to deliver enhanced breast imaging with complementary information for an improved diagnostic process.

  4. A comparison of the performance of modern screen-film and digital mammography systems.

    PubMed

    Monnin, P; Gutierrez, D; Bulling, S; Lepori, D; Valley, J-F; Verdun, F R

    2005-06-01

    This work compares the detector performance and image quality of the new Kodak Min-R EV mammography screen-film system with the Fuji CR Profect detector and with other current mammography screen-film systems from Agfa, Fuji and Kodak. Basic image quality parameters (MTF, NPS, NEQ and DQE) were evaluated for a 28 kV Mo/Mo (HVL = 0.646 mm Al) beam using different mAs exposure settings. Compared with other screen-film systems, the new Kodak Min-R EV detector has the highest contrast and a low intrinsic noise level, giving better NEQ and DQE results, especially at high optical density. Thus, the properties of the new mammography film approach those of a fine mammography detector, especially at low frequency range. Screen-film systems provide the best resolution. The presampling MTF of the digital detector has a value of 15% at the Nyquist frequency and, due to the spread size of the laser beam, the use of a smaller pixel size would not permit a significant improvement of the detector resolution. The dual collection reading technology increases significantly the low frequency DQE of the Fuji CR system that can at present compete with the most efficient mammography screen-film systems. PMID:15901958

  5. Elasto-Mammography: Elastic Property Reconstruction in Breast Tissues

    SciTech Connect

    Wang, Z. G.; Liu, Y.; Wang, G.; Sun, L. Z.

    2008-02-15

    Mammography is the primary method for screening and detecting breast cancers. However, it frequently fails to detect small tumors and is not quite specific in terms of tumor benignity and malignancy. The objective of this paper is to develop a new imaging modality called elasto-mammography that generates the modulus elastograms based on conventional mammographs. A new elastic reconstruction method is described based on elastography and mammography for breast tissues. Elastic distribution can be reconstructed through the measurement of displacement provided by mammographic projection. It is shown that the proposed elasto-mammography provides higher sensitivity and specificity than the conventional mammography on its own for breast cancer diagnosis.

  6. Quality metrics for sensor images

    NASA Technical Reports Server (NTRS)

    Ahumada, AL

    1993-01-01

    Methods are needed for evaluating the quality of augmented visual displays (AVID). Computational quality metrics will help summarize, interpolate, and extrapolate the results of human performance tests with displays. The FLM Vision group at NASA Ames has been developing computational models of visual processing and using them to develop computational metrics for similar problems. For example, display modeling systems use metrics for comparing proposed displays, halftoning optimizing methods use metrics to evaluate the difference between the halftone and the original, and image compression methods minimize the predicted visibility of compression artifacts. The visual discrimination models take as input two arbitrary images A and B and compute an estimate of the probability that a human observer will report that A is different from B. If A is an image that one desires to display and B is the actual displayed image, such an estimate can be regarded as an image quality metric reflecting how well B approximates A. There are additional complexities associated with the problem of evaluating the quality of radar and IR enhanced displays for AVID tasks. One important problem is the question of whether intruding obstacles are detectable in such displays. Although the discrimination model can handle detection situations by making B the original image A plus the intrusion, this detection model makes the inappropriate assumption that the observer knows where the intrusion will be. Effects of signal uncertainty need to be added to our models. A pilot needs to make decisions rapidly. The models need to predict not just the probability of a correct decision, but the probability of a correct decision by the time the decision needs to be made. That is, the models need to predict latency as well as accuracy. Luce and Green have generated models for auditory detection latencies. Similar models are needed for visual detection. Most image quality models are designed for static imagery. Watson has been developing a general spatial-temporal vision model to optimize video compression techniques. These models need to be adapted and calibrated for AVID applications.

  7. Estimation of breast percent density in raw and processed full field digital mammography images via adaptive fuzzy c-means clustering and support vector machine segmentation

    SciTech Connect

    Keller, Brad M.; Nathan, Diane L.; Wang Yan; Zheng Yuanjie; Gee, James C.; Conant, Emily F.; Kontos, Despina

    2012-08-15

    Purpose: The amount of fibroglandular tissue content in the breast as estimated mammographically, commonly referred to as breast percent density (PD%), is one of the most significant risk factors for developing breast cancer. Approaches to quantify breast density commonly focus on either semiautomated methods or visual assessment, both of which are highly subjective. Furthermore, most studies published to date investigating computer-aided assessment of breast PD% have been performed using digitized screen-film mammograms, while digital mammography is increasingly replacing screen-film mammography in breast cancer screening protocols. Digital mammography imaging generates two types of images for analysis, raw (i.e., 'FOR PROCESSING') and vendor postprocessed (i.e., 'FOR PRESENTATION'), of which postprocessed images are commonly used in clinical practice. Development of an algorithm which effectively estimates breast PD% in both raw and postprocessed digital mammography images would be beneficial in terms of direct clinical application and retrospective analysis. Methods: This work proposes a new algorithm for fully automated quantification of breast PD% based on adaptive multiclass fuzzy c-means (FCM) clustering and support vector machine (SVM) classification, optimized for the imaging characteristics of both raw and processed digital mammography images as well as for individual patient and image characteristics. Our algorithm first delineates the breast region within the mammogram via an automated thresholding scheme to identify background air followed by a straight line Hough transform to extract the pectoral muscle region. The algorithm then applies adaptive FCM clustering based on an optimal number of clusters derived from image properties of the specific mammogram to subdivide the breast into regions of similar gray-level intensity. Finally, a SVM classifier is trained to identify which clusters within the breast tissue are likely fibroglandular, which are then aggregated into a final dense tissue segmentation that is used to compute breast PD%. Our method is validated on a group of 81 women for whom bilateral, mediolateral oblique, raw and processed screening digital mammograms were available, and agreement is assessed with both continuous and categorical density estimates made by a trained breast-imaging radiologist. Results: Strong association between algorithm-estimated and radiologist-provided breast PD% was detected for both raw (r= 0.82, p < 0.001) and processed (r= 0.85, p < 0.001) digital mammograms on a per-breast basis. Stronger agreement was found when overall breast density was assessed on a per-woman basis for both raw (r= 0.85, p < 0.001) and processed (0.89, p < 0.001) mammograms. Strong agreement between categorical density estimates was also seen (weighted Cohen's {kappa}{>=} 0.79). Repeated measures analysis of variance demonstrated no statistically significant differences between the PD% estimates (p > 0.1) due to either presentation of the image (raw vs processed) or method of PD% assessment (radiologist vs algorithm). Conclusions: The proposed fully automated algorithm was successful in estimating breast percent density from both raw and processed digital mammographic images. Accurate assessment of a woman's breast density is critical in order for the estimate to be incorporated into risk assessment models. These results show promise for the clinical application of the algorithm in quantifying breast density in a repeatable manner, both at time of imaging as well as in retrospective studies.

  8. Positioning challenges in mammography.

    PubMed

    Peart, Olive

    2014-01-01

    Patients presenting for mammography are different ages and sizes and have varying body habitus; they include men, those who arrive on a stretcher or in a wheelchair, and those with very small breasts, large or wide breasts, pectus excavatum or pectus carinatum, a barrel chest, or kyphosis. The true professional must know how to image patients who deviate from the norm. In addition to competent positioning skills and anatomical knowledge, the mammographer needs a thorough knowledge of the various projections and the skills to modify any projection to meet the needs of individual patients. PMID:24614444

  9. A complete software application for automatic registration of x-ray mammography and magnetic resonance images

    SciTech Connect

    Solves-Llorens, J. A.; Rupérez, M. J. Monserrat, C.; Lloret, M.

    2014-08-15

    Purpose: This work presents a complete and automatic software application to aid radiologists in breast cancer diagnosis. The application is a fully automated method that performs a complete registration of magnetic resonance (MR) images and x-ray (XR) images in both directions (from MR to XR and from XR to MR) and for both x-ray mammograms, craniocaudal (CC), and mediolateral oblique (MLO). This new approximation allows radiologists to mark points in the MR images and, without any manual intervention, it provides their corresponding points in both types of XR mammograms and vice versa. Methods: The application automatically segments magnetic resonance images and x-ray images using the C-Means method and the Otsu method, respectively. It compresses the magnetic resonance images in both directions, CC and MLO, using a biomechanical model of the breast that distinguishes the specific biomechanical behavior of each one of its three tissues (skin, fat, and glandular tissue) separately. It makes a projection of both compressions and registers them with the original XR images using affine transformations and nonrigid registration methods. Results: The application has been validated by two expert radiologists. This was carried out through a quantitative validation on 14 data sets in which the Euclidean distance between points marked by the radiologists and the corresponding points obtained by the application were measured. The results showed a mean error of 4.2 ± 1.9 mm for the MRI to CC registration, 4.8 ± 1.3 mm for the MRI to MLO registration, and 4.1 ± 1.3 mm for the CC and MLO to MRI registration. Conclusions: A complete software application that automatically registers XR and MR images of the breast has been implemented. The application permits radiologists to estimate the position of a lesion that is suspected of being a tumor in an imaging modality based on its position in another different modality with a clinically acceptable error. The results show that the application can accelerate the mammographic screening process for high risk populations or for dense breasts.

  10. Near-infrared spectral imaging of the female breast for quantitative oximetry in optical mammography

    SciTech Connect

    Yu Yang; Liu Ning; Sassaroli, Angelo; Fantini, Sergio

    2009-04-01

    We present a hybrid continuous-wave, frequency-domain instrument for near-infrared spectral imaging of the female breast based on a tandem, planar scanning of one illumination optical fiber and one collection optical fiber configured in a transmission geometry. The spatial sampling rate of 25 points/cm{sup 2} is increased to 400 points/cm{sup 2} by postprocessing the data with a 2D cubic spline interpolation. We then apply a previously developed spatial second-derivative algorithm to an edge-corrected intensity image (N-image) to enhance the visibility and resolution of optical inhomogeneities in breast tissue such as blood vessels and tumors. The spectral data at each image pixel consist of 515-point spectra over the 650-900 nm wavelength range, thus featuring a spectral density of two data points per nanometer. We process the measured spectra with a paired-wavelength spectral analysis method to quantify the oxygen saturation of detected optical inhomogeneities, under the assumption that they feature a locally higher hemoglobin concentration. Our initial measurements on two healthy human subjects have generated high-resolution optical mammograms displaying a network of blood vessels with values of hemoglobin saturation typically falling within the 60%-95% range, which is physiologically reasonable. This approach to spectral imaging and oximetry of the breast has the potential to efficiently exploit the high intrinsic contrast provided by hemoglobin in breast tissue and to contribute a useful tool in the detection, diagnosis, and monitoring of breast pathologies.

  11. An interactive method based on the live wire for segmentation of the breast in mammography images.

    PubMed

    Zewei, Zhang; Tianyue, Wang; Li, Guo; Tingting, Wang; Lu, Xu

    2014-01-01

    In order to improve accuracy of computer-aided diagnosis of breast lumps, the authors introduce an improved interactive segmentation method based on Live Wire. This paper presents the Gabor filters and FCM clustering algorithm is introduced to the Live Wire cost function definition. According to the image FCM analysis for image edge enhancement, we eliminate the interference of weak edge and access external features clear segmentation results of breast lumps through improving Live Wire on two cases of breast segmentation data. Compared with the traditional method of image segmentation, experimental results show that the method achieves more accurate segmentation of breast lumps and provides more accurate objective basis on quantitative and qualitative analysis of breast lumps. PMID:25024740

  12. Near-infrared optical mammography with broadband spectral imaging for spatially resolved oximetry

    NASA Astrophysics Data System (ADS)

    Yu, Yang; Sassaroli, Angelo; Homer, Marc J.; Graham, Roger A.; Fantini, Sergio

    2011-02-01

    We report the development of an instrument for diffuse spectral imaging of the human breast operating over the wavelength range 650-900 nm. This instrument images the slightly compressed human breast in a planar geometry by performing a tandem scan, over the x-y plane, of a 3 mm illumination optical fiber and a 5 mm collection optical fiber that are collinear and located on opposite sides of the breast. An edge-correction algorithm accounts for breast thickness variability over the x-y plane, a second-derivative imaging algorithm enhances the display of optical inhomogeneities, and a paired-wavelength spectral method yields oxygenation maps. We report our results of oxygenation mapping in eighteen human subjects, two of which are breast cancer patients, one with a ductal carcinoma in situ, the other with an invasive ductal carcinoma.

  13. Mammographic tissue, breast cancer risk, serial image analysis, and digital mammography. Part 2. Serial breast tissue change and related temporal influences.

    PubMed

    Heine, John J; Malhotra, Poonam

    2002-03-01

    The work presented herein is the second part of a review of breast tissue-related cancer-risk research. Briefly, in part 1, the tissue-risk research is discussed. In this part, factors that influence temporal breast tissue change are reviewed. Since breast composition is correlated with some of the known risk factors, understanding these influences may provide a mechanism for measuring the dynamics of breast cancer risk. The purpose of this work is to provide support for an automated serial mammography study under way at the authors' institution, where the digital mammographic images are acquired with a full-field digital mammography imaging system. At the initiation of the serial study, it was clear that the authors did not fully understand the nature of the problem: automatically comparing similar mammographic scenes acquired at different times. The evidence indicates that there are many factors that influence breast tissue at any given time and thus have the ability to alter the associated radiographic appearance over time. In general, the topics considered herein include aging; involution; breast development; exogenous and endogenous hormonal interactions such as hormone replacement therapy, oral contraceptive use, and menstrual timing; screening sensitivity issues and interval cancers; tumor growth rates; sojourn times; and lifestyle factors such as diet and exercise. Throughout this work, commentaries and suggestive strategies for automated serial image analysis are provided. PMID:11887947

  14. Toward clinical differential phase contrast mammography: preliminary evaluations and image processing schemes

    NASA Astrophysics Data System (ADS)

    Stampanoni, M.; Wang, Z.; Thüring, T.; David, C.; Rössl, E.; van Stevendaal, U.; Köhler, T.; Trippel, M.; Singer, G.; Kubik-Huch, R. A.; Hohl, M. K.; Hauser, N.

    2013-05-01

    Phase contrast and scattering-based X-ray imaging are very promising tools for medical diagnostics because they are able to provide additional and complementary information to traditional absorption-based methods. In this work, we discuss the investigation of three native breast samples with a grating interferometer equipped with a conventional X-ray tube, the full study being published in ref. [1]. We briefly introduce a method to fuse absorption, differential phase and scattering signals into a unique image with improved diagnostic contents. Our approach yields complementary and inaccessible information on the electron density distribution and the small-angle scattering power of the sample which could potentially answer clinically relevant, yet unresolved questions such as the capability to unequivocally discern between (pre-) malignant changes and post-operative scars or to distinguish cancer-invaded regions within healthy tissue.

  15. New mammography screen/film combinations: Imaging characteristics and radiation dose

    SciTech Connect

    Kimme-Smith, C.; Bassett, L.W.; Gold, R.H.; Zheutlin, J.; Gornbein, J.A. )

    1990-04-01

    Five types of film (Kodak OM, Kodak OM-SO177, Konica CM, Dupont Microvision, and Fuji MiMa) exposed in combination with seven different intensifying screens (Min R, Min R Medium, Siemens Orthox MA, Kyokka HR Mammo Fine, Agfa Gevaert Detail S (old and new), and Konica Monarch) were processed for either 90 sec (at 33.3{degrees}C) or 3 min (at 35.0 degrees C). The films imaged a Computerized Imaging Reference System phantom with additional detail test objects placed on its surface to produce four groups of objects with which to evaluate resolution and contrast. For objects that tested resolution, the Kyokka HR Mammo Fine (Fuji) screen was statistically significantly superior; for objects that tested contrast, the Konica Monarch screen was statistically significantly superior. Extended processing did not affect Dupont and Kodak OM film as much as it affected the other films. It did affect contrast for the other films tested. The mean glandular doses from gridless exposures ranged from 32 to 80 mrad (0.32-0.80 mGy) over all film/screen/processing combinations for a 4.5-cm-thick test object. Several new film/screen combinations can provide images superior to the Kodak Min R/OM combination at a reduced radiation dose. The Kyokka HR Mammo Fine (Fuji) screen was found statistically superior in radiographic resolution of mammographic test objects and the Konica Monarch screen was found to be superior in defining contrast.

  16. Signal uniformity of mammography systems and its impact on test results from contrast detail phantoms

    NASA Astrophysics Data System (ADS)

    Kaar, M.; Semturs, F.; Hummel, J.; Hoffmann, R.; Figl, M.

    2015-03-01

    Technical quality assurance (TQA) procedures for mammography systems usually include tests with a contrast-detail phantom. These phantoms contain multiple objects of varying dimensions arranged on a flat body. Exposures of the phantom are then evaluated by an observer, either human or software. One well-known issue of this method is that dose distribution is not uniform across the image area of any mammography system, mainly due to the heel effect. The purpose of this work is to investigate to what extent image quality differs across the detector plane. We analyze a total of 320 homogeneous mammography exposures from 32 radiology institutes. Systems of different models and manufacturers, both computed radiography (CR) and direct radiography (DR) are included. All images were taken from field installations operated within the nationwide Austrian mammography screening program, which includes mandatory continuous TQA. We calculate signal-to-noise ratios (SNR) for 15 regions of interest arranged to cover the area of the phantom. We define the 'signal range' of an image and compare this value categorized by technologies. We found the deviations of SNR greater in anterior-posterior than in lateral direction. SNR ranges are significantly higher for CR systems than for DR systems.

  17. Broadband optical mammography instrument for depth-resolved imaging and local dynamic measurements.

    PubMed

    Krishnamurthy, Nishanth; Kainerstorfer, Jana M; Sassaroli, Angelo; Anderson, Pamela G; Fantini, Sergio

    2016-02-01

    We present a continuous-wave instrument for non-invasive diffuse optical imaging of the breast in a parallel-plate transmission geometry. The instrument measures continuous spectra in the wavelength range 650-1000 nm, with an intensity noise level <1.5% and a spatial sampling rate of 5 points/cm in the x- and y-directions. We collect the optical transmission at four locations, one collinear and three offset with respect to the illumination optical fiber, to recover the depth of optical inhomogeneities in the tissue. We imaged a tissue-like, breast shaped, silicone phantom (6 cm thick) with two embedded absorbing structures: a black circle (1.7 cm in diameter) and a black stripe (3 mm wide), designed to mimic a tumor and a blood vessel, respectively. The use of a spatially multiplexed detection scheme allows for the generation of on-axis and off-axis projection images simultaneously, as opposed to requiring multiple scans, thus decreasing scan-time and motion artifacts. This technique localizes detected inhomogeneities in 3D and accurately assigns their depth to within 1 mm in the ideal conditions of otherwise homogeneous tissue-like phantoms. We also measured induced hemodynamic changes in the breast of a healthy human subject at a selected location (no scanning). We applied a cyclic, arterial blood pressure perturbation by alternating inflation (to a pressure of 200 mmHg) and deflation of a pneumatic cuff around the subject's thigh at a frequency of 0.05 Hz, and measured oscillations with amplitudes up to 1 μM and 0.2 μM in the tissue concentrations of oxyhemoglobin and deoxyhemoglobin, respectively. These hemodynamic oscillations provide information about the vascular structure and functional integrity in tissue, and may be used to assess healthy or abnormal perfusion in a clinical setting. PMID:26931870

  18. Broadband optical mammography instrument for depth-resolved imaging and local dynamic measurements

    NASA Astrophysics Data System (ADS)

    Krishnamurthy, Nishanth; Kainerstorfer, Jana M.; Sassaroli, Angelo; Anderson, Pamela G.; Fantini, Sergio

    2016-02-01

    We present a continuous-wave instrument for non-invasive diffuse optical imaging of the breast in a parallel-plate transmission geometry. The instrument measures continuous spectra in the wavelength range 650-1000 nm, with an intensity noise level <1.5% and a spatial sampling rate of 5 points/cm in the x- and y-directions. We collect the optical transmission at four locations, one collinear and three offset with respect to the illumination optical fiber, to recover the depth of optical inhomogeneities in the tissue. We imaged a tissue-like, breast shaped, silicone phantom (6 cm thick) with two embedded absorbing structures: a black circle (1.7 cm in diameter) and a black stripe (3 mm wide), designed to mimic a tumor and a blood vessel, respectively. The use of a spatially multiplexed detection scheme allows for the generation of on-axis and off-axis projection images simultaneously, as opposed to requiring multiple scans, thus decreasing scan-time and motion artifacts. This technique localizes detected inhomogeneities in 3D and accurately assigns their depth to within 1 mm in the ideal conditions of otherwise homogeneous tissue-like phantoms. We also measured induced hemodynamic changes in the breast of a healthy human subject at a selected location (no scanning). We applied a cyclic, arterial blood pressure perturbation by alternating inflation (to a pressure of 200 mmHg) and deflation of a pneumatic cuff around the subject's thigh at a frequency of 0.05 Hz, and measured oscillations with amplitudes up to 1 μM and 0.2 μM in the tissue concentrations of oxyhemoglobin and deoxyhemoglobin, respectively. These hemodynamic oscillations provide information about the vascular structure and functional integrity in tissue, and may be used to assess healthy or abnormal perfusion in a clinical setting.

  19. The use of a figure-of-merit (FOM) for optimisation in digital mammography: a literature review.

    PubMed

    Borg, M; Badr, I; Royle, G J

    2012-08-01

    The use of image quality parameters in digital mammography such as contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR) and detective quantum efficiency (DQE) has been widespread, with the intention of detector evaluation and/or quantitative evaluation of the system performance. These parameters are useful in ensuring adequate system performance when tests are done against international standards or guidelines. Parameters like CNR are relative quantities that lie within a range that is manufacturer and system dependent. The use of a figure-of-merit (FOM) is a relatively new concept as a tool in digital mammography permitting quantitative assessment in terms of image quality and patient dose. This review summarises the available evidence for the use and applicability of an FOM in digital mammography. PMID:22232780

  20. The life span of silicone gel breast implants and a comparison of mammography, ultrasonography, and magnetic resonance imaging in detecting implant rupture: a meta-analysis.

    PubMed

    Goodman, C M; Cohen, V; Thornby, J; Netscher, D

    1998-12-01

    Because of the growing concern surrounding the integrity and life span of silicone gel breast implants and the reported variations in the diagnostic accuracy of various imaging techniques in identifying ruptured implants, the authors undertook a meta-analysis of articles in the scientific literature to examine these concerns. They were able to include reports from the literature that detailed the condition and removal of 1,099 breast implants during the past 7 years. The median life span of a silicone gel implant was estimated to be 16.4 years. Of the implants, 79.1% were intact at 10 years, falling to 48.7% by 15 years. The sensitivities and specificities of three imaging modalities used in the diagnosis of implant rupture (mammography, ultrasonography, and magnetic resonance imaging [MRI]) were also evaluated and compared statistically in an effort to discover which of the three techniques might serve as the most reliable screening tool in the diagnosis of gel implant rupture. The sensitivity of mammography for finding a ruptured implant is 28.4% with a specificity of 92.9%. Ultrasonography has a sensitivity and specificity of 59.0% and 76.8% respectively compared with MRI, which was 78.1% and 80.0% respectively. For implants in place for 10 years, one would need to image 3.3 implants by ultrasound to identify a single possible rupture. However, because of the 76.8% specificity, 8.1 implants would need to be imaged to find a confirmed intraoperative rupture. This was similar to MRI, in which 3.1 implants would need to be imaged to detect one suspected rupture, and 6.1 implants would need to be imaged to find one intraoperatively confirmed rupture. The authors do not recommend either ultrasound or MRI as a screening tool based on their meta-analysis. PMID:9869129

  1. Dual-energy contrast-enhanced mammography.

    PubMed

    Travieso Aja, M M; Rodríguez Rodríguez, M; Alayón Hernández, S; Vega Benítez, V; Luzardo, O P

    2014-01-01

    The degree of vascularization in breast lesions is related to their malignancy. For this reason, functional diagnostic imaging techniques have become important in recent years. Dual-energy contrast-enhanced mammography is a new, apparently promising technique in breast cancer that provides information about the degree of vascularization of the lesion in addition to the morphological information provided by conventional mammography. This article describes the state of the art for dual-energy contrast-enhanced mammography. Based on 15 months' clinical experience, we illustrate this review with clinical cases that allow us to discuss the advantages and limitations of this technique. PMID:25086679

  2. A comparison of the performance of digital mammography systems.

    PubMed

    Monnin, P; Gutierrez, D; Bulling, S; Guntern, D; Verdun, F R

    2007-03-01

    An objective analysis of image quality parameters was performed for six digital mammography systems. The presampled modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE) for the systems were determined at different doses, for 28 kVp with a Mo/Mo or W/Al target/filter combination and 2 mm of additional aluminium filtration. The flat-panel units have higher MTF and DQE in the mid to high frequency range than standard CR systems. The highest DQE, over the whole dose range, is for the slit-scanning direct photon counting system. Dual-side read CR can overcome the inherent x-ray absorption and signal collection limitations of standard CR mammography, improving the low-frequency DQE by 40%, to the same level as full-field systems, but it does not improve the poor spatial resolution of phosphor. PMID:17441236

  3. A comparison of the performance of digital mammography systems

    SciTech Connect

    Monnin, P.; Gutierrez, D.; Bulling, S.; Guntern, D.; Verdun, F. R.

    2007-03-15

    An objective analysis of image quality parameters was performed for six digital mammography systems. The presampled modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE) for the systems were determined at different doses, for 28 kVp with a Mo/Mo or W/Al target/filter combination and 2 mm of additional aluminium filtration. The flat-panel units have higher MTF and DQE in the mid to high frequency range than standard CR systems. The highest DQE, over the whole dose range, is for the slit-scanning direct photon counting system. Dual-side read CR can overcome the inherent x-ray absorption and signal collection limitations of standard CR mammography, improving the low-frequency DQE by 40%, to the same level as full-field systems, but it does not improve the poor spatial resolution of phosphor.

  4. Study on computer-aided diagnosis of hepatic MR imaging and mammography

    SciTech Connect

    Zhang Xuejun

    2005-04-01

    It is well known that the liver is an organ easily attacked by diseases. The purpose of this study is to develop a computer-aided diagnosis (CAD) scheme for helping radiologists to differentiate hepatic diseases more efficiently. Our software named LIVERANN integrated the magnetic resonance (MR) imaging findings with different pulse sequences to classify the five categories of hepatic diseases by using the artificial neural network (ANN) method. The intensity and homogeneity within the region of interest (ROI) delineated by a radiologist were automatically calculated to obtain numerical data by the program for input signals to the ANN. Outputs were the five pathological categories of hepatic diseases (hepatic cyst, hepatocellular carcinoma, dysplasia in cirrhosis, cavernous hemangioma, and metastasis). The experiment demonstrated a testing accuracy of 93% from 80 patients. In order to differentiate the cirrhosis from normal liver, the volume ratio of left to whole (LTW) was proposed to quantify the degree of cirrhosis by three-dimensional (3D) volume analysis. The liver region was firstly extracted from computed tomography (CT) or MR slices based on edge detection algorithms, and then separated into left lobe and right lobe by the hepatic umbilical fissure. The volume ratio of these two parts showed that the LTW ratio in the liver was significantly improved in the differentiation performance, with (25.6%{+-}4.3%) in cirrhosis versus the normal liver (16.4%{+-}5.4%). In addition, the application of the ANN method for detecting clustered microcalcifications in masses on mammograms was described here as well. A new structural ANN, so-called a shift-invariant artificial neural network (SIANN), was integrated with our triple-ring filter (TRF) method in our CAD system. As the result, the sensitivity of detecting clusters was improved from 90% by our previous TRF method to 95% by using both SIANN and TRF.

  5. Infrared image quality evaluation method without reference image

    NASA Astrophysics Data System (ADS)

    Yue, Song; Ren, Tingting; Wang, Chengsheng; Lei, Bo; Zhang, Zhijie

    2013-09-01

    Since infrared image quality depends on many factors such as optical performance and electrical noise of thermal imager, image quality evaluation becomes an important issue which can conduce to both image processing afterward and capability improving of thermal imager. There are two ways of infrared image quality evaluation, with or without reference image. For real-time thermal image, the method without reference image is preferred because it is difficult to get a standard image. Although there are various kinds of methods for evaluation, there is no general metric for image quality evaluation. This paper introduces a novel method to evaluate infrared image without reference image from five aspects: noise, clarity, information volume and levels, information in frequency domain and the capability of automatic target recognition. Generally, the basic image quality is obtained from the first four aspects, and the quality of target is acquired from the last aspect. The proposed method is tested on several infrared images captured by different thermal imagers. Calculate the indicators and compare with human vision results. The evaluation shows that this method successfully describes the characteristics of infrared image and the result is consistent with human vision system.

  6. Objective Quality Assessment of Interpolated Natural Images.

    PubMed

    Yeganeh, Hojatollah; Rostami, Mohammad; Wang, Zhou

    2015-11-01

    Image interpolation techniques that create high-resolution images from low-resolution (LR) images are widely used in real world applications, but how to evaluate the quality of interpolated images is not a well-resolved issue. Subjective assessment methods are useful and reliable, but are also slow and expensive. Here, we propose an objective method to assess the quality of an interpolated natural image using the available LR image as a reference. Our method adopts a natural scene statistics (NSS) framework, where image quality degradation is gauged by the deviation of its statistical features from the NSS models trained upon high-quality natural images. Two distortion measures are proposed, namely, interpolated natural image distortion (IND) and weighted IND. Validations by subjective tests show that the proposed approach performs statistically equivalent or sometimes better than an average human subject. Moreover, we demonstrate the potential application of the proposed method in parameter tuning of image interpolation algorithms. PMID:26186792

  7. Breast Imaging Reporting and Data System (BI-RADS) breast composition descriptors: Automated measurement development for full field digital mammography

    SciTech Connect

    Fowler, E. E.; Sellers, T. A.; Lu, B.; Heine, J. J.

    2013-11-15

    Purpose: The Breast Imaging Reporting and Data System (BI-RADS) breast composition descriptors are used for standardized mammographic reporting and are assessed visually. This reporting is clinically relevant because breast composition can impact mammographic sensitivity and is a breast cancer risk factor. New techniques are presented and evaluated for generating automated BI-RADS breast composition descriptors using both raw and calibrated full field digital mammography (FFDM) image data.Methods: A matched case-control dataset with FFDM images was used to develop three automated measures for the BI-RADS breast composition descriptors. Histograms of each calibrated mammogram in the percent glandular (pg) representation were processed to create the new BR{sub pg} measure. Two previously validated measures of breast density derived from calibrated and raw mammograms were converted to the new BR{sub vc} and BR{sub vr} measures, respectively. These three measures were compared with the radiologist-reported BI-RADS compositions assessments from the patient records. The authors used two optimization strategies with differential evolution to create these measures: method-1 used breast cancer status; and method-2 matched the reported BI-RADS descriptors. Weighted kappa (κ) analysis was used to assess the agreement between the new measures and the reported measures. Each measure's association with breast cancer was evaluated with odds ratios (ORs) adjusted for body mass index, breast area, and menopausal status. ORs were estimated as per unit increase with 95% confidence intervals.Results: The three BI-RADS measures generated by method-1 had κ between 0.25–0.34. These measures were significantly associated with breast cancer status in the adjusted models: (a) OR = 1.87 (1.34, 2.59) for BR{sub pg}; (b) OR = 1.93 (1.36, 2.74) for BR{sub vc}; and (c) OR = 1.37 (1.05, 1.80) for BR{sub vr}. The measures generated by method-2 had κ between 0.42–0.45. Two of these measures were significantly associated with breast cancer status in the adjusted models: (a) OR = 1.95 (1.24, 3.09) for BR{sub pg}; (b) OR = 1.42 (0.87, 2.32) for BR{sub vc}; and (c) OR = 2.13 (1.22, 3.72) for BR{sub vr}. The radiologist-reported measures from the patient records showed a similar association, OR = 1.49 (0.99, 2.24), although only borderline statistically significant.Conclusions: A general framework was developed and validated for converting calibrated mammograms and continuous measures of breast density to fully automated approximations for the BI-RADS breast composition descriptors. The techniques are general and suitable for a broad range of clinical and research applications.

  8. Univariant assessment of the quality of images

    NASA Astrophysics Data System (ADS)

    Jung, Mathieu; Leger, Dominique; Gazalet, Marc G.

    2002-07-01

    To evaluate the quality of images, most methods compare a degraded image to a perfect reference. Nevertheless in many cases, a reference does not exist. We propose an original univariant (i.e., without a reference) method based on the use of artificial neural networks. The principle behind it is to first teach a neural network to assess image quality using images taken from a pool of known examples, then use it to assess the quality of unknown images. The defects considered are compression artifacts, ringing, local singularities, etc. To simplify, only images with defects that are not mixed with one another were first used. Two illustrative examples are presented: assessment of the quality of JPEG compressed images and detection of local defects. The quality of the images is assessed without a reference and with error less than 6% - 7% compared to the bivariant method that was learned. Our method can even be used to model some very simple visual comportment.

  9. An Underwater Color Image Quality Evaluation Metric.

    PubMed

    Yang, Miao; Sowmya, Arcot

    2015-12-01

    Quality evaluation of underwater images is a key goal of underwater video image retrieval and intelligent processing. To date, no metric has been proposed for underwater color image quality evaluation (UCIQE). The special absorption and scattering characteristics of the water medium do not allow direct application of natural color image quality metrics especially to different underwater environments. In this paper, subjective testing for underwater image quality has been organized. The statistical distribution of the underwater image pixels in the CIELab color space related to subjective evaluation indicates the sharpness and colorful factors correlate well with subjective image quality perception. Based on these, a new UCIQE metric, which is a linear combination of chroma, saturation, and contrast, is proposed to quantify the non-uniform color cast, blurring, and low-contrast that characterize underwater engineering and monitoring images. Experiments are conducted to illustrate the performance of the proposed UCIQE metric and its capability to measure the underwater image enhancement results. They show that the proposed metric has comparable performance to the leading natural color image quality metrics and the underwater grayscale image quality metrics available in the literature, and can predict with higher accuracy the relative amount of degradation with similar image content in underwater environments. Importantly, UCIQE is a simple and fast solution for real-time underwater video processing. The effectiveness of the presented measure is also demonstrated by subjective evaluation. The results show better correlation between the UCIQE and the subjective mean opinion score. PMID:26513783

  10. Contrast to Noise Ratio and Contrast Detail Analysis in Mammography:A Monte Carlo Study

    NASA Astrophysics Data System (ADS)

    Metaxas, V.; Delis, H.; Kalogeropoulou, C.; Zampakis, P.; Panayiotakis, G.

    2015-09-01

    The mammographic spectrum is one of the major factors affecting image quality in mammography. In this study, a Monte Carlo (MC) simulation model was used to evaluate image quality characteristics of various mammographic spectra. The anode/filter combinations evaluated, were those traditionally used in mammography, for tube voltages between 26 and 30 kVp. The imaging performance was investigated in terms of Contrast to Noise Ratio (CNR) and Contrast Detail (CD) analysis, by involving human observers, utilizing a mathematical CD phantom. Soft spectra provided the best characteristics in terms of both CNR and CD scores, while tube voltage had a limited effect. W-anode spectra filtered with k-edge filters demonstrated an improved performance, that sometimes was better compared to softer x-ray spectra, produced by Mo or Rh anode. Regarding the filter material, k-edge filters showed superior performance compared to Al filters.

  11. Quality Metrics Evaluation of Hyperspectral Images

    NASA Astrophysics Data System (ADS)

    Singh, A. K.; Kumar, H. V.; Kadambi, G. R.; Kishore, J. K.; Shuttleworth, J.; Manikandan, J.

    2014-11-01

    In this paper, the quality metrics evaluation on hyperspectral images has been presented using k-means clustering and segmentation. After classification the assessment of similarity between original image and classified image is achieved by measurements of image quality parameters. Experiments were carried out on four different types of hyperspectral images. Aerial and spaceborne hyperspectral images with different spectral and geometric resolutions were considered for quality metrics evaluation. Principal Component Analysis (PCA) has been applied to reduce the dimensionality of hyperspectral data. PCA was ultimately used for reducing the number of effective variables resulting in reduced complexity in processing. In case of ordinary images a human viewer plays an important role in quality evaluation. Hyperspectral data are generally processed by automatic algorithms and hence cannot be viewed directly by human viewers. Therefore evaluating quality of classified image becomes even more significant. An elaborate comparison is made between k-means clustering and segmentation for all the images by taking Peak Signal-to-Noise Ratio (PSNR), Mean Square Error (MSE), Maximum Squared Error, ratio of squared norms called L2RAT and Entropy. First four parameters are calculated by comparing the quality of original hyperspectral image and classified image. Entropy is a measure of uncertainty or randomness which is calculated for classified image. Proposed methodology can be used for assessing the performance of any hyperspectral image classification techniques.

  12. Dual-energy imaging using a digital scanned multi-slit system for mammography: evaluation of a differential beam filtering technique

    NASA Astrophysics Data System (ADS)

    Bornefalk, Hans; Hemmendorff, Magnus; Hjärn, Torbjörn

    2006-03-01

    This paper describes a method for single exposure contrast-enhanced dual-energy imaging of tumors utilizing a scanned multi-slit system for digital mammography. This photon counting system employs an array of silicon strip detectors in an edge-on geometry. In the multi-slit setup, the line detectors and pre-collimator slits are aligned orthogonal to the scan direction. This geometry is advantageous to dual-energy imaging, since it allows differential filtering of the x-ray beam in the pre-collimator slits. A high-energy image is constructed from those lines where the filter material has been chosen to harden the x-ray beam and the low-energy image from the lines with a filter producing softer beams. Both images are obtained in the same scan, eliminating the need to change tube voltages and anode materials and minimizing the risk of motion artifacts. The method is illustrated on a purpose-built phantom and logarithmic subtraction of the images produces images essentially free of anatomical clutter with the contrast-enhanced targets clearly visible.

  13. Process perspective on image quality evaluation

    NASA Astrophysics Data System (ADS)

    Leisti, Tuomas; Halonen, Raisa; Kokkonen, Anna; Weckman, Hanna; Mettänen, Marja; Lensu, Lasse; Ritala, Risto; Oittinen, Pirkko; Nyman, Göte

    2008-01-01

    The psychological complexity of multivariate image quality evaluation makes it difficult to develop general image quality metrics. Quality evaluation includes several mental processes and ignoring these processes and the use of a few test images can lead to biased results. By using a qualitative/quantitative (Interpretation Based Quality, IBQ) methodology, we examined the process of pair-wise comparison in a setting, where the quality of the images printed by laser printer on different paper grades was evaluated. Test image consisted of a picture of a table covered with several objects. Three other images were also used, photographs of a woman, cityscape and countryside. In addition to the pair-wise comparisons, observers (N=10) were interviewed about the subjective quality attributes they used in making their quality decisions. An examination of the individual pair-wise comparisons revealed serious inconsistencies in observers' evaluations on the test image content, but not on other contexts. The qualitative analysis showed that this inconsistency was due to the observers' focus of attention. The lack of easily recognizable context in the test image may have contributed to this inconsistency. To obtain reliable knowledge of the effect of image context or attention on subjective image quality, a qualitative methodology is needed.

  14. Comparative performance of modern digital mammography systems in a large breast screening program

    SciTech Connect

    Yaffe, Martin J. Bloomquist, Aili K.; Hunter, David M.; Mawdsley, Gordon E.; Chiarelli, Anna M.; Muradali, Derek; Mainprize, James G.

    2013-12-15

    Purpose: To compare physical measures pertaining to image quality among digital mammography systems utilized in a large breast screening program. To examine qualitatively differences in these measures and differences in clinical cancer detection rates between CR and DR among sites within that program. Methods: As part of the routine quality assurance program for screening, field measurements are made of several variables considered to correlate with the diagnostic quality of medical images including: modulation transfer function, noise equivalent quanta, d′ (an index of lesion detectability) and air kerma to allow estimation of mean glandular dose. In addition, images of the mammography accreditation phantom are evaluated. Results: It was found that overall there were marked differences between the performance measures of DR and CR mammography systems. In particular, the modulation transfer functions obtained with the DR systems were found to be higher, even for larger detector element sizes. Similarly, the noise equivalent quanta, d′, and the phantom scores were higher, while the failure rates associated with low signal-to-noise ratio and high dose were lower with DR. These results were consistent with previous findings in the authors’ program that the breast cancer detection rates at sites employing CR technology were, on average, 30.6% lower than those that used DR mammography. Conclusions: While the clinical study was not large enough to allow a statistically powered system-by-system assessment of cancer detection accuracy, the physical measures expressing spatial resolution, and signal-to-noise ratio are consistent with the published finding that sites employing CR systems had lower cancer detection rates than those using DR systems for screening mammography.

  15. Disparities in Screening Mammography Services by Race/Ethnicity and Health Insurance

    PubMed Central

    Allgood, Kristi L.; Whitman, Steve; Conant, Emily

    2012-01-01

    Abstract Background Black and Hispanic women are diagnosed at a later stage of breast cancer than white women. Differential access to specialists, diffusion of technology, and affiliation with an academic medical center may be related to this stage disparity. Methods We analyzed data from a mammography facility survey for the metropolitan region of Chicago, Illinois, to assess in part whether quality breast imaging services were equally accessed by non-Hispanic white, non-Hispanic black, and Hispanic women and by women with and without private insurance. Of 49 screening facilities within the city of Chicago, 43 facilities completed the survey, and 40 facilities representing about 149,000 mammograms, including all major academic facilities, provided data on patient race/ethnicity. Results Among women receiving mammograms at the facilities we studied, white women were more likely than black or Hispanic women to have mammograms at academic facilities, at facilities that relied exclusively on breast imaging specialists to read mammograms, and at facilities where digital mammography was available (p<0.001). Women with private insurance were similarly more likely than women without private insurance to have mammograms at facilities with these characteristics (p<0.001). Conclusions Black and Hispanic women and women without private insurance are more likely than white women and women with private insurance to obtain mammography screening at facilities with less favorable characteristics. A disparity in use of high-quality mammography may be contributing to disparities in breast cancer mortality. PMID:21942866

  16. JPEG2000 still image coding quality.

    PubMed

    Chen, Tzong-Jer; Lin, Sheng-Chieh; Lin, You-Chen; Cheng, Ren-Gui; Lin, Li-Hui; Wu, Wei

    2013-10-01

    This work demonstrates the image qualities between two popular JPEG2000 programs. Two medical image compression algorithms are both coded using JPEG2000, but they are different regarding the interface, convenience, speed of computation, and their characteristic options influenced by the encoder, quantization, tiling, etc. The differences in image quality and compression ratio are also affected by the modality and compression algorithm implementation. Do they provide the same quality? The qualities of compressed medical images from two image compression programs named Apollo and JJ2000 were evaluated extensively using objective metrics. These algorithms were applied to three medical image modalities at various compression ratios ranging from 10:1 to 100:1. Following that, the quality of the reconstructed images was evaluated using five objective metrics. The Spearman rank correlation coefficients were measured under every metric in the two programs. We found that JJ2000 and Apollo exhibited indistinguishable image quality for all images evaluated using the above five metrics (r > 0.98, p < 0.001). It can be concluded that the image quality of the JJ2000 and Apollo algorithms is statistically equivalent for medical image compression. PMID:23589187

  17. Perspectives of Mobile Versus Fixed Mammography in Santa Clara County, California: A Focus Group Study

    PubMed Central

    Chang-Halpenny, Christine; Kumarasamy, Narmadan A; Venegas, Angela; Braddock III, Clarence H

    2016-01-01

    Objective: Our aim was to examine underserved women’s perceptions on mobile versus fixed mammography in Santa Clara, California through a focus group study. Background: Research has shown that medically underserved women have higher breast cancer mortality rates correlated with under-screening and a disproportional rate of late-stage diagnosis. The Community Health Partnership in Santa Clara County, California runs the Community Mammography Access Project (CMAP) that targets nearly 20,000 medically underserved women over the age of 40 in the county through the collaborative effort of an existing safety net of healthcare providers. However, little data exists on the advantages or disadvantages of mobile mammography units from the patient perspective.  Methods: We assessed underserved women’s perspectives on mammography services in Santa Clara County through two focus groups from women screened at mobile or fixed site programs. Patients were recruited from both CMAP clinics and a county hospital, and focus group data were analyzed using content analysis. Results: We found that women from both the mobile and fixed sites shared similar motivating factors for getting a mammogram. Both groups recognized that screening was uncomfortable but necessary for good health and had positive feedback about their personal physicians. However, mobile participants, in particular, appreciated the atmosphere of mobile screening, reported shorter wait times, and remarked on the good communication from the clinic staff and empathetic treatment they received. However, mobile participants also expressed concern about the quality of films at mobile sites due to delayed initial reading of the films.  Conclusions: Mobile mammography offers a unique opportunity for women of underserved populations to access high satisfaction screenings, and it encourages a model similar to CMAP in other underserved areas. However, emphasis should be placed on providing a warm and welcoming environment for patients and ensuring the quality of mammography images.  PMID:27014528

  18. Breast Density in Mammography and Magnetic Resonance Imaging in High Risk Women and Women with Breast Cancer

    PubMed Central

    Albert, Marissa; Schnabel, Freya; Chun, Jennifer; Schwartz, Shira; Lee, Jiyon; Leite, Ana Paula Klautau; Moy, Linda

    2015-01-01

    Structured Abstract Purpose To evaluate the relationship between mammographic breast density (MBD), background parenchymal enhancement (BPE), and fibroglandular tissue (FGT) in women with breast cancer (BC) and at high risk for developing BC. Methods Our institutional database was queried for patients who underwent mammography and MRI. Results 403 (85%) had BC and 72 (15%) were at high risk. MBD (p=0.0005), BPE (p<0.0001), and FGT (p=0.02) were all higher in high risk women compared to the BC group. Conclusions Higher levels of MBD, BPE and FGT are seen in women at higher risk for developing BC when compared to women with BC. PMID:26351036

  19. Composite x-ray image assembly for large-field digital mammography with one- and two-dimensional positioning of a focal plane array

    NASA Technical Reports Server (NTRS)

    Halama, G.; McAdoo, J.; Liu, H.

    1998-01-01

    To demonstrate the feasibility of a novel large-field digital mammography technique, a 1024 x 1024 pixel Loral charge-coupled device (CCD) focal plane array (FPA) was positioned in a mammographic field with one- and two-dimensional scan sequences to obtain 950 x 1800 pixel and 3600 x 3600 pixel composite images, respectively. These experiments verify that precise positioning of FPAs produced seamless composites and that the CCD mosaic concept has potential for high-resolution, large-field imaging. The proposed CCD mosaic concept resembles a checkerboard pattern with spacing left between the CCDs for the driver and readout electronics. To obtain a complete x-ray image, the mosaic must be repositioned four times, with an x-ray exposure at each position. To reduce the patient dose, a lead shield with appropriately patterned holes is placed between the x-ray source and the patient. The high-precision motorized translation stages and the fiber-coupled-scintillating-screen-CCD sensor assembly were placed in the position usually occupied by the film cassette. Because of the high mechanical precision, seamless composites were constructed from the subimages. This paper discusses the positioning, image alignment procedure, and composite image results. The paper only addresses the formation of a seamless composite image from subimages and will not consider the effects of the lead shield, multiple CCDs, or the speed of motion.

  20. Cognitive issues in image quality measurement

    NASA Astrophysics Data System (ADS)

    de Ridder, Huib

    2001-01-01

    Designers of imaging systems, image processing algorithms, etc., usually take for granted that methods for assessing perceived image quality produce unbiased estimates of the viewers' quality impression. Quality judgments, however, are affected by the judgment strategies induced by the experimental procedures. In this paper the results of two experiments are presented illustrating the influence judgment strategies can have on quality judgments. The first experiment concerns contextual effects due to the composition of the stimulus sets. Subjects assessed the sharpness of two differently composed sets of blurred versions of one static image. The sharpness judgments for the blurred images present in both stimulus sets were found to be dependent on the composition of the set as well as the scaling technique employed. In the second experiment subjects assessed either the overall quality or the overall impairment of manipulated and standard JPEG-coded images containing two main artifacts. The results indicate a systematic different between the quality and impairment judgments that could be interpreted as instruction-based different weighting of the two artifacts. Again, some influence of scaling techniques was observed. The results of both experiments underscore the important role judgment strategies play in the psychophysical evaluation of image quality. Ignoring this influence on quality judgments may lead to invalid conclusions about the viewers' impression of image quality.

  1. Retinal image quality assessment using generic features

    NASA Astrophysics Data System (ADS)

    Fasih, Mahnaz; Langlois, J. M. Pierre; Ben Tahar, Houssem; Cheriet, Farida

    2014-03-01

    Retinal image quality assessment is an important step in automated eye disease diagnosis. Diagnosis accuracy is highly dependent on the quality of retinal images, because poor image quality might prevent the observation of significant eye features and disease manifestations. A robust algorithm is therefore required in order to evaluate the quality of images in a large database. We developed an algorithm for retinal image quality assessment based on generic features that is independent from segmentation methods. It exploits the local sharpness and texture features by applying the cumulative probability of blur detection metric and run-length encoding algorithm, respectively. The quality features are combined to evaluate the image's suitability for diagnosis purposes. Based on the recommendations of medical experts and our experience, we compared a global and a local approach. A support vector machine with radial basis functions was used as a nonlinear classifier in order to classify images to gradable and ungradable groups. We applied our methodology to 65 images of size 2592×1944 pixels that had been graded by a medical expert. The expert evaluated 38 images as gradable and 27 as ungradable. The results indicate very good agreement between the proposed algorithm's predictions and the medical expert's judgment: the sensitivity and specificity for the local approach are respectively 92% and 94%. The algorithm demonstrates sufficient robustness to identify relevant images for automated diagnosis.

  2. Image quality assessment based on distortion identification

    NASA Astrophysics Data System (ADS)

    Chetouani, Aladine; Beghdadi, Azeddine

    2011-01-01

    A New Global Full-Reference Image Quality System based on classification and fusion scheme is proposed. It consists of many steps. The first step is devoted to the identification of the type of degradation contained in a given image based a Linear Discriminant Analysis (LDA) classifier using some common Image Quality Metric (IQM) as feature inputs. An IQM per degradation (IQM-D) is then used to estimate the quality of the image. For a given degradation type, the appropriate IQM-D is derived by combining the top three best IQMs using an Artificial Neural Network model. The performance of the proposed scheme is evaluated first in terms of good degradation identification. Then, for each distortion type the image quality estimation is evaluated in terms of good correlation with the subjective judgments using the TID 2008 image database.

  3. Technology assessment: observer study directly compares screen/film to CR mammography

    NASA Astrophysics Data System (ADS)

    Fletcher-Heath, Lynn; Richards, Anne; Ryan-Kron, Susan

    2007-03-01

    A new study supports and expands upon a previous reporting that computed radiography (CR) mammography offers as good, or better, image quality than state-of-the-art screen/film mammography. The suitability of CR mammography is explored through qualitative and quantitative study components: feature comparison and cancer detection rates of each modality. Images were collected from 150 normal and 50 biopsy-confirmed subjects representing a range of breast and pathology types. Comparison views were collected without releasing compression, using automatic exposure control on Kodak MIN-R films, followed by CR. Digital images were displayed as both softcopy (S/C) and hardcopy (H/C) for the feature comparison, and S/C for the cancer detection task. The qualitative assessment used preference scores from five board-certified radiologists obtained while viewing 100 screen/film-CR pairs from the cancer subjects for S/C and H/C CR output. Fifteen general image-quality features were rated, and up to 12 additional features were rated for each pair, based on the pathology present. Results demonstrate that CR is equivalent or preferred to conventional mammography for overall image quality (89% S/C, 95% H/C), image contrast (95% S/C, 98% H/C), sharpness (86% S/C, 93% H/C), and noise (94% S/C, 91% H/C). The quantitative objective was satisfied by asking 10 board-certified radiologists to provide a BI-RADS TM score and probability of malignancy per breast for each modality of the 200 cases. At least 28 days passed between observations of the same case. Average sensitivity and specificity was 0.89 and 0.82 for CR and 0.91 and 0.82 for screen/film, respectively.

  4. Analysis of SAR images quality degradation factors

    NASA Astrophysics Data System (ADS)

    Dawidowicz, Bartek; Samczynski, Piotr; Smolarczyk, Maciej; Kuzmiuk, Marek

    2006-03-01

    Quality of Synthetic Aperture Radar (SAR) images is very sensitive for both pulse signal spectrum purity and stability of the radar platform flight path. In this paper detailed influence of this factors on SAR images quality is analyzed. Analysis is based mainly on computer simulations. This kind of analysis is very important because not all SAR image degradation factors can be compensated using digital signal processing algorithms.

  5. Effect of the glandular composition on digital breast tomosynthesis image quality and dose optimisation.

    PubMed

    Marques, T; Ribeiro, A; Di Maria, S; Belchior, A; Cardoso, J; Matela, N; Oliveira, N; Janeiro, L; Almeida, P; Vaz, P

    2015-07-01

    In the image quality assessment for digital breast tomosynthesis (DBT), a breast phantom with an average percentage of 50 % glandular tissue is seldom used, which may not be representative of the breast tissue composition of the women undergoing such examination. This work aims at studying the effect of the glandular composition of the breast on the image quality taking into consideration different sizes of lesions. Monte Carlo simulations were performed using the state-of-the-art computer program PENELOPE to validate the image acquisition system of the DBT equipment as well as to calculate the mean glandular dose for each projection image and for different breast compositions. The integrated PENELOPE imaging tool (PenEasy) was used to calculate, in mammography, for each clinical detection task the X-ray energy that maximises the figure of merit. All the 2D cranial-caudal projections for DBT were simulated and then underwent the reconstruction process applying the Simultaneous Algebraic Reconstruction Technique. Finally, through signal-to-noise ratio analysis, the image quality in DBT was assessed. PMID:25836692

  6. Validation of a digital mammographic unit model for an objective and highly automated clinical image quality assessment.

    PubMed

    Perez-Ponce, Hector; Daul, Christian; Wolf, Didier; Noel, Alain

    2013-08-01

    In mammography, image quality assessment has to be directly related to breast cancer indicator (e.g. microcalcifications) detectability. Recently, we proposed an X-ray source/digital detector (XRS/DD) model leading to such an assessment. This model simulates very realistic contrast-detail phantom (CDMAM) images leading to gold disc (representing microcalcifications) detectability thresholds that are very close to those of real images taken under the simulated acquisition conditions. The detection step was performed with a mathematical observer. The aim of this contribution is to include human observers into the disc detection process in real and virtual images to validate the simulation framework based on the XRS/DD model. Mathematical criteria (contrast-detail curves, image quality factor, etc.) are used to assess and to compare, from the statistical point of view, the cancer indicator detectability in real and virtual images. The quantitative results given in this paper show that the images simulated by the XRS/DD model are useful for image quality assessment in the case of all studied exposure conditions using either human or automated scoring. Also, this paper confirms that with the XRS/DD model the image quality assessment can be automated and the whole time of the procedure can be drastically reduced. Compared to standard quality assessment methods, the number of images to be acquired is divided by a factor of eight. PMID:23207102

  7. The Future of Mammography: Radiology Residents’ Experiences, Attitudes, and Opinions

    PubMed Central

    Baxi, Shrujal S.; Snow, Jacqueline G.; Liberman, Laura; Elkin, Elena B.

    2011-01-01

    OBJECTIVE The objective of our study was to assess the experiences and preferences of radiology residents with respect to breast imaging. MATERIALS AND METHODS We surveyed radiology residents at 26 programs in New York and New Jersey. Survey topics included plans for subspecialty training, beliefs, and attitudes toward breast imaging and breast cancer screening and the likelihood of interpreting mammography in the future. RESULTS Three hundred forty-four residents completed the survey (response rate, 62%). The length of time spent training in breast imaging varied from no dedicated time (37%) to 1–8 weeks (40%) to more than 9 weeks (23%). Most respondents (97%) agreed that mammography is important to women’s health. More than 85% of residents believed that mammography should be interpreted by breast imaging specialists. Respondents shared negative views about mammography, agreeing with statements that the field was associated with a high risk of malpractice (99%), stress (94%), and low reimbursement (68%). Respondents endorsed several positive attributes of mammography, including job availability (97%), flexible work schedules (94%), and few calls or emergencies (93%). Most radiology residents (93%) said that they were likely to pursue subspecialty training, and 7% expressed interest in breast imaging fellowships. CONCLUSION Radiology residents’ negative and positive views about mammography seem to be independent of time spent training in mammography and of future plans to pursue fellowship training in breast imaging. Systematic assessment of the plans and preferences of radiology residents can facilitate the development of strategies to attract trainees to careers in breast imaging. PMID:20489113

  8. Optimal photon energy comparison between digital breast tomosynthesis and mammography: a case study.

    PubMed

    Di Maria, S; Baptista, M; Felix, M; Oliveira, N; Matela, N; Janeiro, L; Vaz, P; Orvalho, L; Silva, A

    2014-06-01

    A comparison, in terms of the optimal energy that maximizes the image quality between digital breast tomosynthesis (DBT) and digital mammography (DM) was performed in a MAMMOMAT Inspiration system (Siemens) based on amorphous selenium flat panel detector. In this paper we measured the image quality by the signal difference-to-noise ratio (SDNR), and the patient risk by the mean glandular dose (MGD). Using these quantities we compared the optimal voltage that maximizes the image quality both in breast tomosynthesis and standard mammography acquisition mode. The comparison for the two acquisition modes was performed for a W/Rh anode filter combinations by using a 4.5 cm tissue equivalent mammography phantom. Moreover, in order to check if the used equipment was quantum noise limited, the relation of the relative noise with respect to the detector dose was evaluated. Results showed that in the tomosynthesis acquisition mode the optimal voltage is 28 kV, whereas in standard mammography the optimal voltage is 30 kV. The automatic exposure control (AEC) of the system selects 28 kV as optimal voltage both for DBT and DM. Monte Carlo simulations showed a qualitative agreement with the AEC selection system, since an optimal monochromatic energy of 20 keV was found both for DBT and DM. Moreover, the check about the noise showed that the system is not completely quantum noise limited, and this issue could explain the experimental slight difference in terms of optimal voltage between DBT and DM. According to these results, the use of higher voltage settings is not justified for the improvement of the image quality during a DBT examination. PMID:24613514

  9. Does image quality matter? Impact of resolution and noise on mammographic task performance

    SciTech Connect

    Saunders, Robert S. Jr.; Baker, Jay A.; Delong, David M.; Johnson, Jeff P.; Samei, Ehsan

    2007-10-15

    The purpose of this study was to examine the effects of different resolution and noise levels on task performance in digital mammography. This study created an image set with images at three different resolution levels, corresponding to three digital display devices, and three different noise levels, with noise magnitudes similar to full clinical dose, half clinical dose, and quarter clinical dose. The images were read by five experienced breast imaging radiologists. The data were then analyzed to compute two accuracy statistics (overall classification accuracy and lesion detection accuracy) and performance at four diagnostic tasks (detection of microcalcifications, benign masses, malignant masses, and discrimination of benign and malignant masses). Human observer results showed decreasing display resolution had little effect on overall classification accuracy and individual diagnostic task performance, but increasing noise caused overall classification accuracy to decrease by a statistically significant 21% as the breast dose went to one quarter of its normal clinical value. The noise effects were most prominent for the tasks of microcalcification detection and mass discrimination. When the noise changed from full clinical dose to quarter clinical dose, the microcalcification detection performance fell from 89% to 67% and the mass discrimination performance decreased from 93% to 79%, while malignant mass detection performance remained relatively constant with values of 88% and 84%, respectively. As a secondary aim, the image set was also analyzed by two observer models to examine whether their performance was similar to humans. Observer models differed from human observers and each other in their sensitivity to resolution degradation and noise. The primary conclusions of this study suggest that quantum noise appears to be the dominant image quality factor in digital mammography, affecting radiologist performance much more profoundly than display resolution.

  10. Automatic quality assessment of planetary images

    NASA Astrophysics Data System (ADS)

    Sidiropoulos, P.; Muller, J.-P.

    2015-10-01

    A significant fraction of planetary images are corrupted beyond the point that much scientific meaning can be extracted. For example, transmission errors result in missing data which is unrecoverable. The available planetary image datasets include many such "bad data", which both occupy valuable scientific storage resources and create false impressions about planetary image availability for specific planetary objects or target areas. In this work, we demonstrate a pipeline that we have developed to automatically assess the quality of planetary images. Additionally, this method discriminates between different types of image degradation, such as low-quality originating from camera flaws or low-quality triggered by atmospheric conditions, etc. Examples of quality assessment results for Viking Orbiter imagery will be also presented.

  11. Impact of SAR image quality on recognition

    NASA Astrophysics Data System (ADS)

    Carlson, Daniel W.; Montagnino, Lee J.; Frankot, Robert T.

    2005-05-01

    Automatic target recognition (ATR) performance is a function of image quality and its representation in the signature model generation and used in the ATR training process. This paper reports ATR performance as a function of synthetic aperture radar (SAR) image quality parameters including clutter-to-noise ratio (CNR) and multiplicative noise ratio (MNR). Images with specified image quality values were produced by introducing controlled degradations to the MSTAR public release data. Two different families of ATR algorithms, the statistical model-based classifier of DeVore, et al., and optimal tradeoff synthetic discriminant function (OTSDF) are applied to those data. Target classification accuracy was measured as a function of CNR/MNR for both the training and test data, indicating sensitivity of performance to a priori knowledge of these particular image quality parameters. Confusion matrices are expanded to include target aspect bins, providing visibility into performance as a function of aspect angle.

  12. Clinical utility of positron emission mammography.

    PubMed

    Glass, Shannon B; Shah, Zeeshan A

    2013-07-01

    Several imaging modalities have been introduced over recent years to better screen for and stage breast cancer. Positron emission mammography (PEM) has been approved by the US Food and Drug Administration and introduced into clinical use as a diagnostic adjunct to mammography and breast ultrasonography. PEM has higher resolution and a more localized field of view than positron emission tomography-computed tomography and can be performed on patients to stage a newly diagnosed malignancy. Review of mammograms together with magnetic resonance or PEM images improves detection of disease. PMID:23814402

  13. Configuration of automatic exposure control on mammography units for computed radiography to match patient dose of screen film systems

    NASA Astrophysics Data System (ADS)

    Yang, Chang-Ying Joseph; Huang, Weidong

    2009-02-01

    Computed radiography (CR) is considered a drop-in addition or replacement for traditional screen-film (SF) systems in digital mammography. Unlike other technologies, CR has the advantage of being compatible with existing mammography units. One of the challenges, however, is to properly configure the automatic exposure control (AEC) on existing mammography units for CR use. Unlike analogue systems, the capture and display of digital CR images is decoupled. The function of AEC is changed from ensuring proper and consistent optical density of the captured image on film to balancing image quality with patient dose needed for CR. One of the preferences when acquiring CR images under AEC is to use the same patient dose as SF systems. The challenge is whether the existing AEC design and calibration process-most of them proprietary from the X-ray systems manufacturers and tailored specifically for SF response properties-can be adapted for CR cassettes, in order to compensate for their response and attenuation differences. This paper describes the methods for configuring the AEC of three different mammography units models to match the patient dose used for CR with those that are used for a KODAK MIN-R 2000 SF System. Based on phantom test results, these methods provide the dose level under AEC for the CR systems to match with the dose of SF systems. These methods can be used in clinical environments that require the acquisition of CR images under AEC at the same dose levels as those used for SF systems.

  14. Radiation protection program for early detection of breast cancer in a mammography facility

    SciTech Connect

    Mariana, Villagomez Casimiro E-mail: cesar@fisica.unam.mx; Cesar, Ruiz Trejo E-mail: cesar@fisica.unam.mx; Ruby, Espejo Fonseca

    2014-11-07

    Mammography is the best tool for early detection of Breast Cancer. In this diagnostic radiology modality it is necessary to establish the criteria to ensure the proper use and operation of the equipment used to obtain mammographic images in order to contribute to the safe use of ionizing radiation. The aim of the work was to implement at FUCAM-AC the radiation protection program which must be established for patients and radiation workers according to Mexican standards [1–4]. To achieve this goal, radiation protection and quality control manuals were elaborated [5]. Furthermore, a quality control program (QCP) in the mammography systems (analog/digital), darkroom included, has been implemented. Daily sensitometry, non-variability of the image quality, visualizing artifacts, revision of the equipment mechanical stability, compression force and analysis of repetition studies are some of the QCP routine tests that must be performed by radiological technicians of this institution as a set of actions to ensure the protection of patients. Image quality and patients dose assessment were performed on 4 analog equipment installed in 2 mobile units. In relation to dose assessment, all equipment passed the acceptance criteria (<3 mGy per projection). The image quality test showed that most images (70%)– presented artifacts. A brief summary of the results of quality control tests applied to the equipment and film processor are presented. To maintain an adequate level of quality and safety at FUCAM-AC is necessary that the proposed radiation protection program in this work is applied.

  15. Radiation protection program for early detection of breast cancer in a mammography facility

    NASA Astrophysics Data System (ADS)

    Villagomez Casimiro, Mariana; Ruiz Trejo, Cesar; Espejo Fonseca, Ruby

    2014-11-01

    Mammography is the best tool for early detection of Breast Cancer. In this diagnostic radiology modality it is necessary to establish the criteria to ensure the proper use and operation of the equipment used to obtain mammographic images in order to contribute to the safe use of ionizing radiation. The aim of the work was to implement at FUCAM-AC the radiation protection program which must be established for patients and radiation workers according to Mexican standards [1-4]. To achieve this goal, radiation protection and quality control manuals were elaborated [5]. Furthermore, a quality control program (QCP) in the mammography systems (analog/digital), darkroom included, has been implemented. Daily sensitometry, non-variability of the image quality, visualizing artifacts, revision of the equipment mechanical stability, compression force and analysis of repetition studies are some of the QCP routine tests that must be performed by radiological technicians of this institution as a set of actions to ensure the protection of patients. Image quality and patients dose assessment were performed on 4 analog equipment installed in 2 mobile units. In relation to dose assessment, all equipment passed the acceptance criteria (<3 mGy per projection). The image quality test showed that most images (70%)- presented artifacts. A brief summary of the results of quality control tests applied to the equipment and film processor are presented. To maintain an adequate level of quality and safety at FUCAM-AC is necessary that the proposed radiation protection program in this work is applied.

  16. Perceptual image quality: Effects of tone characteristics

    PubMed Central

    Delahunt, Peter B.; Zhang, Xuemei; Brainard, David H.

    2007-01-01

    Tone mapping refers to the conversion of luminance values recorded by a digital camera or other acquisition device, to the luminance levels available from an output device, such as a monitor or a printer. Tone mapping can improve the appearance of rendered images. Although there are a variety of algorithms available, there is little information about the image tone characteristics that produce pleasing images. We devised an experiment where preferences for images with different tone characteristics were measured. The results indicate that there is a systematic relation between image tone characteristics and perceptual image quality for images containing faces. For these images, a mean face luminance level of 46–49 CIELAB L* units and a luminance standard deviation (taken over the whole image) of 18 CIELAB L* units produced the best renderings. This information is relevant for the design of tone-mapping algorithms, particularly as many images taken by digital camera users include faces. PMID:17235365

  17. Development of contrast digital mammography.

    PubMed

    Skarpathiotakis, Mia; Yaffe, Martin J; Bloomquist, Aili K; Rico, Dan; Muller, Serge; Rick, Andreas; Jeunehomme, Fanny

    2002-10-01

    Development of breast tumors is often accompanied by angiogenesis--the formation of new blood vessels. It is possible to image the effects of this process by tracking the uptake and washout of contrast agents in the vicinity of a lesion. In this article, a method for carrying out contrast subtraction mammography on a full-field digital mammography unit is described. Spectral measurements and modeling were performed to optimize the choice of x-ray target, kilovoltage and x-ray beam filtration for contrast digital mammography (CDM) on an available digital mammography system. Phantom studies were carried out to determine the sensitivity of CDM to iodine. Detection of iodine area densities of 0.3 mg/cm2 is possible for a circular object with a radius of 1.3 mm, which allows detection of uptake levels in the breast typically seen with cancer and some benign breast conditions. It was found that with a molybdenum anode x-ray tube, copper filtration could be used to effectively shape the x-ray spectrum to maximize the proportion of x rays with energies above the k edge of iodine. Simple logarithmic subtraction was found to be adequate in suppressing background signals dependent on the x-ray beam intensity and background thickness of the breast. The total x-ray dose from the procedure ranges between 1 and 3 mGy, similar to that from a conventional single view film mammogram. A clinical pilot study is currently being carried out to evaluate this technique. PMID:12408316

  18. Asymmetric scatter kernels for software-based scatter correction of gridless mammography

    NASA Astrophysics Data System (ADS)

    Wang, Adam; Shapiro, Edward; Yoon, Sungwon; Ganguly, Arundhuti; Proano, Cesar; Colbeth, Rick; Lehto, Erkki; Star-Lack, Josh

    2015-03-01

    Scattered radiation remains one of the primary challenges for digital mammography, resulting in decreased image contrast and visualization of key features. While anti-scatter grids are commonly used to reduce scattered radiation in digital mammography, they are an incomplete solution that can add radiation dose, cost, and complexity. Instead, a software-based scatter correction method utilizing asymmetric scatter kernels is developed and evaluated in this work, which improves upon conventional symmetric kernels by adapting to local variations in object thickness and attenuation that result from the heterogeneous nature of breast tissue. This fast adaptive scatter kernel superposition (fASKS) method was applied to mammography by generating scatter kernels specific to the object size, x-ray energy, and system geometry of the projection data. The method was first validated with Monte Carlo simulation of a statistically-defined digital breast phantom, which was followed by initial validation on phantom studies conducted on a clinical mammography system. Results from the Monte Carlo simulation demonstrate excellent agreement between the estimated and true scatter signal, resulting in accurate scatter correction and recovery of 87% of the image contrast originally lost to scatter. Additionally, the asymmetric kernel provided more accurate scatter correction than the conventional symmetric kernel, especially at the edge of the breast. Results from the phantom studies on a clinical system further validate the ability of the asymmetric kernel correction method to accurately subtract the scatter signal and improve image quality. In conclusion, software-based scatter correction for mammography is a promising alternative to hardware-based approaches such as anti-scatter grids.

  19. Image quality and wafer level optics

    NASA Astrophysics Data System (ADS)

    Dagan, Y.; Humpston, G.

    2010-05-01

    Increasing demand from consumers to integrate camera modules into electronic devices, such as cell phones, has driven the cost of camera modules down very rapidly. Now that most cell phones include at least one camera, consumers are starting to ask for better image quality - without compromising on the cost. Wafer level optics has emerged over the past few years as an innovative technology enabling simultaneous manufacturing of thousands of lenses, at the wafer level. Using reflow-compatible materials to manufacture these lenses permits a reduction in the cost and size of camera module, thus answering the market demand for lowering the cost. But what about image quality? The author will present image quality analysis that was conducted for both VGA and megapixel camera resolutions. Comparison between conventional camera modules and wafer level camera modules shows wafer level technology brings equivalent, if not better, image quality performance compared to conventional camera modules.

  20. Disparities in Screening Mammography

    PubMed Central

    Peek, Monica E; Han, Jini H

    2004-01-01

    OBJECTIVE This paper describes trends in screening mammography utilization over the past decade and assesses the remaining disparities in mammography use among medically underserved women. We also describe the barriers to mammography and report effective interventions to enhance utilization. DESIGN We reviewed medline and other databases as well as relevant bibliographies. MAIN RESULTS The United States has dramatically improved its use of screening mammography over the past decade, with increased rates observed in every demographic group. Disparities in screening mammography are decreasing among medically underserved populations but still persist among racial/ethnic minorities and low-income women. Additionally, uninsured women and those with no usual care have the lowest rates of reported mammogram use. However, despite apparent increases in mammogram utilization, there is growing evidence that limitations in the national survey databases lead to overestimations of mammogram use, particularly among low-income racial and ethnic minorities. CONCLUSIONS The United States may be farther from its national goals of screening mammography, particularly among underserved women, than current data suggests. We should continue to support those interventions that increase mammography use among the medically underserved by addressing the barriers such as cost, language and acculturation limitations, deficits in knowledge and cultural beliefs, literacy and health system barriers such as insurance and having a source regular of medical care. Addressing disparities in the diagnostic and cancer treatment process should also be a priority in order to affect significant change in health outcomes among the underserved. PMID:15009798

  1. Optical Image Quality and the Cone Mosaic

    NASA Astrophysics Data System (ADS)

    Snyder, Allan W.; Bossomaier, Terry R. J.; Hughes, Austin

    1986-01-01

    Contrary to the orthodox view that optical image quality should ``match'' the photoreceptor grain, anatomical data from the eyes of various animals suggest that the image quality is significantly superior to the potential resolution of the cone mosaic in most retinal regions. A new theory is presented to explain the existence of this relation and to better appreciate eye design. It predicts that photoreceptors are potentially visible through the natural optics.

  2. Toward clinically relevant standardization of image quality.

    PubMed

    Samei, Ehsan; Rowberg, Alan; Avraham, Ellie; Cornelius, Craig

    2004-12-01

    In recent years, notable progress has been made on standardization of medical image presentations in the definition and implementation of the Digital Imaging and Communications in Medicine (DICOM) Grayscale Standard Display Function (GSDF). In parallel, the American Association of Physicists in Medicine (AAPM) Task Group 18 has provided much needed guidelines and tools for visual and quantitative assessment of medical display quality. In spite of these advances, however, there are still notable gaps in the effectiveness of DICOM GSDF to assure consistent and high-quality display of medical images. In additions the degree of correlation between display technical data and diagnostic usability and performance of displays remains unclear. This article proposes three specific steps that DICOM, AAPM, and ACR may collectively take to bridge the gap between technical performance and clinical use: (1) DICOM does not provide means and acceptance criteria to evaluate the conformance of a display device to GSDF or to address other image quality characteristics. DICOM can expand beyond luminance response, extending the measurable, quantifiable elements of TG18 such as reflection and resolution. (2) In a large picture archiving and communication system (PACS) installation, it is critical to continually track the appropriate use and performance of multiple display devices. DICOM may help with this task by adding a Device Service Class to the standard to provide for communication and control of image quality parameters between applications and devices, (3) The question of clinical significance of image quality metrics has rarely been addressed by prior efforts. In cooperation with AAPM, the American College of Radiology (ACR), and the Society for Computer Applications in Radiology (SCAR), DICOM may help to initiate research that will determine the clinical consequence of variations in image quality metrics (eg, GSDF conformance) and to define what constitutes image quality from a diagnostic perspective. Implementation of these three initiatives may further the reach and impact of DICOM toward quality medicine. PMID:15551103

  3. Tradeoffs between image quality and dose.

    PubMed

    Seibert, J Anthony

    2004-10-01

    Image quality takes on different perspectives and meanings when associated with the concept of as low as reasonably achievable (ALARA), which is chiefly focused on radiation dose delivered as a result of a medical imaging procedure. ALARA is important because of the increased radiosensitivity of children to ionizing radiation and the desire to keep the radiation dose low. By the same token, however, image quality is also important because of the need to provide the necessary information in a radiograph in order to make an accurate diagnosis. Thus, there are tradeoffs to be considered between image quality and radiation dose, which is the main topic of this article. ALARA does not necessarily mean the lowest radiation dose, nor, when implemented, does it result in the least desirable radiographic images. With the recent widespread implementation of digital radiographic detectors and displays, a new level of flexibility and complexity confronts the technologist, physicist, and radiologist in optimizing the pediatric radiography exam. This is due to the separation of the acquisition, display, and archiving events that were previously combined by the screen-film detector, which allows for compensation for under- and overexposures, image processing, and on-line image manipulation. As explained in the article, different concepts must be introduced for a better understanding of the tradeoffs encountered when dealing with digital radiography and ALARA. In addition, there are many instances during the image acquisition/display/interpretation process in which image quality and associated dose can be compromised. This requires continuous diligence to quality control and feedback mechanisms to verify that the goals of image quality, dose and ALARA are achieved. PMID:15558260

  4. Ultrasound mammography

    NASA Astrophysics Data System (ADS)

    Mensah, Serge; Franceschini, Emilie; Pauzin, Marie-Christine

    2007-02-01

    We introduce a near-field formulation of the acoustic field scattered by a soft tissue organ. This derivation is based on the Huygens-Fresnel principle that describes the scattered field as the result of the interferential scheme of all the secondary spherical waves. This leads us to define a new Fourier transform which yields a spectrum whose harmonic components have an elliptical spatial support. Based on these projections, we define an Elliptical Radon transform that enables us to reconstruct either the impedance or the celerity maps of an acoustical model characterized in terms of impedance and celerity fluctuations. The formulation is very similar to that developed in the far-field domain where the Radon transform pair is derived from an harmonic plane wave decomposition. This formulation allows us to introduce the Ductal Tomography, following the example of the Ductal Echography, that provides a systematic inspection of each mammary lobe, in order to reveal breast lesions at an early stage. In order to review the performances obtained with current echographs in view of specific experiment (numerical simulations), we develop a computer phantom that gains in realism. This 2-D anatomical phantom is an axial cut of the ductolobular structure corresponding to a daisy-like internal arrangement with petals (lobes) radiating around the nipple, for healthy and pathological situations. The different constitutive tissues and ducts are characterized in terms of density and celerity parameters whose spatial distributions are defined with specific random density laws. The use of a velocity-pressure formulation permits us to model time domain acoustic wave propagation. Broadband US pulses are transmitted and measured in diffraction around the breast with a ring antenna, the images are reconstructed using the elliptical back-projection-based procedure mentioned above.

  5. A comparative study of volumetric breast density estimation in digital mammography and magnetic resonance imaging: results from a high-risk population

    NASA Astrophysics Data System (ADS)

    Kontos, Despina; Xing, Ye; Bakic, Predrag R.; Conant, Emily F.; Maidment, Andrew D. A.

    2010-03-01

    We performed a study to compare methods for volumetric breast density estimation in digital mammography (DM) and magnetic resonance imaging (MRI) for a high-risk population of women. DM and MRI images of the unaffected breast from 32 women with recently detected abnormalities and/or previously diagnosed breast cancer (age range 31-78 yrs, mean 50.3 yrs) were retrospectively analyzed. DM images were analyzed using QuantraTM (Hologic Inc). The MRI images were analyzed using a fuzzy-C-means segmentation algorithm on the T1 map. Both methods were compared to Cumulus (Univ. Toronto). Volumetric breast density estimates from DM and MRI are highly correlated (r=0.90, p<=0.001). The correlation between the volumetric and the area-based density measures is lower and depends on the training background of the Cumulus software user (r=0.73-84, p<=0.001). In terms of absolute values, MRI provides the lowest volumetric estimates (mean=14.63%), followed by the DM volumetric (mean=22.72%) and area-based measures (mean=29.35%). The MRI estimates of the fibroglandular volume are statistically significantly lower than the DM estimates for women with very low-density breasts (p<=0.001). We attribute these differences to potential partial volume effects in MRI and differences in the computational aspects of the image analysis methods in MRI and DM. The good correlation between the volumetric and the area-based measures, shown to correlate with breast cancer risk, suggests that both DM and MRI volumetric breast density measures can aid in breast cancer risk assessment. Further work is underway to fully-investigate the association between volumetric breast density measures and breast cancer risk.

  6. Combination of digital mammography with semi-automated 3D breast ultrasound.

    PubMed

    Kapur, Ajay; Carson, Paul L; Eberhard, Jeffrey; Goodsitt, Mitchell M; Thomenius, Kai; Lokhandwalla, Murtuza; Buckley, Donald; Roubidoux, Marilyn A; Helvie, Mark A; Booi, Rebecca C; LeCarpentier, Gerald L; Erkamp, Ramon Q; Chan, Heang-Ping; Fowlkes, J Brian; Thomas, Jerry A; Landberg, Cynthia E

    2004-08-01

    This paper describes work aimed at combining 3D ultrasound with full-field digital mammography via a semi-automatic prototype ultrasound scanning mechanism attached to the digital mammography system gantry. Initial efforts to obtain high x-ray and ultrasound image quality through a compression paddle are proving successful. Registration between the x-ray mammogram and ultrasound image volumes is quite promising when the breast is stably compressed. This prototype system takes advantage of many synergies between the co-registered digital mammography and pulse-echo ultrasound image data used for breast cancer detection and diagnosis. In addition, innovative combinations of advanced US and X-ray applications are being implemented and tested along with the basic modes. The basic and advanced applications are those that should provide relatively independent information about the breast tissues. Advanced applications include x-ray tomosynthesis, for 3D delineation of mammographic structures, and non-linear elasticity and 3D color flow imaging by ultrasound, for mechanical and physiological information unavailable from conventional, non-contrast x-ray and ultrasound imaging. PMID:15270583

  7. Technology transfer in digital mammography. Report of the Joint National Cancer Institute-National Aeronautics and Space Administration workshop of May 19-20, 1993.

    PubMed

    Winfield, D; Silbiger, M; Brown, G S; Clarke, L; Dwyer, S; Yaffe, M; Shtern, F

    1994-04-01

    Digital mammography is one of the most promising novel technologies for further improvement of early detection of breast cancer, offering important potential advantages: 1) improved image quality; 2) digital image processing for improved lesion contrast; 3) computer-aided diagnosis for enhanced radiologic interpretation; and 4) teleradiology for facilitated radiologic consultation. The Diagnostic Imaging Research Branch of the National Cancer Institute (NCI) recently funded an international, multidisciplinary, multi-institutional Digital Mammography Development Group for collaborations between NCI, the academic community, and industry to facilitate the integrated development and implementation of digital mammographic systems. Currently, however, digital mammography faces a number of fundamental technological roadblocks: 1) cost-effective digital detectors and displays for imaging systems; 2) the need for novel algorithms for image processing and computer-aided diagnosis; and 3) high performance, low cost digital networks to provide an "information superhighway" for teleradiology. To solve some of these technological problems, the Diagnostic Imaging Research Branch of NCI joined efforts with the Technology Transfer Division of the National Aeronautics and Space Administration to pursue a federal technology transfer program in digital mammography. The authors discuss the findings and recommendations of the workshop entitled "Technology Transfer in Digital Mammography," which was organized and held jointly by the NCI and the National Aeronautics and Space Administration in May, 1993. Numerous innovative technologies of varying degree of promise for digital mammography were presented at the conference. In this article, specific technologies presented at the workshop by the federal and federally-supported laboratories are described, and critiques of these technologies by the leaders of the medical imaging community are presented. PMID:7913457

  8. Normalized Noise Power Spectrum of Full Field Digital Mammography System

    NASA Astrophysics Data System (ADS)

    Isa, Norriza Mohd; Wan Hassan, Wan Muhamad Saridan

    2010-01-01

    A method to measure noise power spectrum of a full field digital mammography system is presented. The effect of X-ray radiation dose, size and configuration of region of interest on normalized noise power spectrum (NNPS) was investigated. Flat field images were acquired using RQA-M2 beam quality technique (Mo/Mo anode-filter, 28 kV, 2 mm Al) with different clinical radiation doses. The images were cropped at about 4 cm from the edge of the breast wall and then divided into different size of non-overlapping or overlapping segments. NNPS was determined through detrending, 2-D fast Fourier transformation and normalization. Our measurement shows that high radiation dose gave lower NNPS at a specific beam quality.

  9. Normalized Noise Power Spectrum of Full Field Digital Mammography System

    SciTech Connect

    Isa, Norriza Mohd; Wan Hassan, Wan Muhamad Saridan

    2010-01-05

    A method to measure noise power spectrum of a full field digital mammography system is presented. The effect of X-ray radiation dose, size and configuration of region of interest on normalized noise power spectrum (NNPS) was investigated. Flat field images were acquired using RQA-M2 beam quality technique (Mo/Mo anode-filter, 28 kV, 2 mm Al) with different clinical radiation doses. The images were cropped at about 4 cm from the edge of the breast wall and then divided into different size of non-overlapping or overlapping segments. NNPS was determined through detrending, 2-D fast Fourier transformation and normalization. Our measurement shows that high radiation dose gave lower NNPS at a specific beam quality.

  10. Color image processing for date quality evaluation

    NASA Astrophysics Data System (ADS)

    Lee, Dah Jye; Archibald, James K.

    2010-01-01

    Many agricultural non-contact visual inspection applications use color image processing techniques because color is often a good indicator of product quality. Color evaluation is an essential step in the processing and inventory control of fruits and vegetables that directly affects profitability. Most color spaces such as RGB and HSV represent colors with three-dimensional data, which makes using color image processing a challenging task. Since most agricultural applications only require analysis on a predefined set or range of colors, mapping these relevant colors to a small number of indexes allows simple and efficient color image processing for quality evaluation. This paper presents a simple but efficient color mapping and image processing technique that is designed specifically for real-time quality evaluation of Medjool dates. In contrast with more complex color image processing techniques, the proposed color mapping method makes it easy for a human operator to specify and adjust color-preference settings for different color groups representing distinct quality levels. Using this color mapping technique, the color image is first converted to a color map that has one color index represents a color value for each pixel. Fruit maturity level is evaluated based on these color indices. A skin lamination threshold is then determined based on the fruit surface characteristics. This adaptive threshold is used to detect delaminated fruit skin and hence determine the fruit quality. The performance of this robust color grading technique has been used for real-time Medjool date grading.

  11. Developing a Comprehensive Database Management System for Organization and Evaluation of Mammography Datasets

    PubMed Central

    Wu, Yirong; Rubin, Daniel L; Woods, Ryan W; Elezaby, Mai; Burnside, Elizabeth S

    2014-01-01

    We aimed to design and develop a comprehensive mammography database system (CMDB) to collect clinical datasets for outcome assessment and development of decision support tools. A Health Insurance Portability and Accountability Act (HIPAA) compliant CMDB was created to store multi-relational datasets of demographic risk factors and mammogram results using the Breast Imaging Reporting and Data System (BI-RADS) lexicon. The CMDB collected both biopsy pathology outcomes, in a breast pathology lexicon compiled by extending BI-RADS, and our institutional breast cancer registry. The audit results derived from the CMDB were in accordance with Mammography Quality Standards Act (MQSA) audits and national benchmarks. The CMDB has managed the challenges of multi-level organization demanded by the complexity of mammography practice and lexicon development in pathology. We foresee that the CMDB will be useful for efficient quality assurance audits and development of decision support tools to improve breast cancer diagnosis. Our procedure of developing the CMDB provides a framework to build a detailed data repository for breast imaging quality control and research, which has the potential to augment existing resources. PMID:25368510

  12. Perceptual image quality and telescope performance ranking

    NASA Astrophysics Data System (ADS)

    Lentz, Joshua K.; Harvey, James E.; Marshall, Kenneth H.; Salg, Joseph; Houston, Joseph B.

    2010-08-01

    Launch Vehicle Imaging Telescopes (LVIT) are expensive, high quality devices intended for improving the safety of vehicle personnel, ground support, civilians, and physical assets during launch activities. If allowed to degrade from the combination of wear, environmental factors, and ineffective or inadequate maintenance, these devices lose their ability to provide adequate quality imagery to analysts to prevent catastrophic events such as the NASA Space Shuttle, Challenger, accident in 1986 and the Columbia disaster of 2003. A software tool incorporating aberrations and diffraction that was developed for maintenance evaluation and modeling of telescope imagery is presented. This tool provides MTF-based image quality metric outputs which are correlated to ascent imagery analysts' perception of image quality, allowing a prediction of usefulness of imagery which would be produced by a telescope under different simulated conditions.

  13. Web-based Mammography Audit Feedback

    PubMed Central

    Geller, BM; Ichikawa, L; Miglioretti, DL; Eastman, D

    2013-01-01

    Introduction Interpreting screening mammography accurately is challenging and requires ongoing education to maintain and improve interpretative skills. Recognizing this, many countries with organized breast screening programs have developed audit and feedback systems based on their national performance data to help radiologists assess and improve their skills. We developed and pilot tested an interactive website to provide screening and diagnostic mammography audit feedback with comparisons to national and regional benchmarks. Methods and Materials Radiologists who participate in three Breast Cancer Surveillance Consortium registries in the United States were invited during 2009 and 2010 to use a website that provides tabular and graphical displays of mammography audit reports with comparisons to national and regional performance measures. We collected data on the use of and perceptions of the website. Results Thirty-five of 111 invited radiologists used the web site from 1–5 times in a year. The most popular measure was sensitivity for both screening and diagnostic mammography while a table with all measures was the most visited page. Of the 13 radiologist who completed the post use survey, all found it easy to use and navigate, 11 found the benchmarks useful, and 9 reported that they intend to improve a specific outcome measure this year. Conclusions An interactive website to provide customized mammography audit feedback reports to radiologists has the potential to be a powerful tool in improving interpretive performance. The conceptual framework of customized audit feedback reports can also be generalized to other imaging tests. PMID:22623571

  14. Visual quality assessment of watermarked medical images

    NASA Astrophysics Data System (ADS)

    Dowling, Jason A.; Planitz, Birgit M.; Maeder, Anthony J.; Du, Jiang; Pham, Binh; Boyd, Colin; Chen, Shaokang; Bradley, Andrew P.; Crozier, Stuart

    2007-03-01

    Increasing transmission of medical images across multiple user systems raises concerns for image security. Hiding watermark information in medical image data files is one solution for enhancing security and privacy protection of data. Medical image watermarking however is not a widely studied area, due partially to speculations on loss in viewer performance caused by degradation of image information. Such concerns are addressed if the amount of information lost due to watermarking can be kept at minimal levels and below visual perception thresholds. This paper describes experiments where three alternative visual quality metrics were used to assess the degradation caused by watermarking medical images. Magnetic Resonance Imaging (MRI) and Computed Tomography (CT) medical images were watermarked using different methods: Block based Discrete Cosine Transform (DCT) and Discrete Wavelet Transform (DWT) with various embedding strengths. The visual degradation of each watermarking parameter setting was assessed using Peak Signal-to-Noise Ratio (PSNR), Structural Similarity Measure (SSIM) and Steerable Visual Difference Predictor (SVDP) numerical metrics. The suitability of each of the three numerical metrics for medical image watermarking visual quality assessment is noted. In addition, subjective test results from human observers are used to suggest visual degradation thresholds.

  15. Optimization of the exposure parameters with signal-to-noise ratios considering human visual characteristics in digital mammography

    NASA Astrophysics Data System (ADS)

    Yamada, Maki; Kato, Yuri; Fujita, Naotoshi; Kodera, Yoshie

    2010-04-01

    The use of digital mammography systems has become widespread recently. However, the optimal exposure parameters are uncertain in clinical practice. We need to optimize the exposure parameter in digital mammography while maximizing image quality and minimizing patient dose. The purpose of this study was to evaluate the most beneficial exposure variable-tube voltage for each compressed breast thickness-with these indices: noise power spectrum, noise equivalent quanta, detective quantum efficiency, and signal-to-noise ratios (SNR). In this study, the SNRs were derived from the perceived statistical decision theory model with the internal noise of eye-brain system (SNRi), contrived and studied by Loo LN1), Ishida M et al. 2) These image quality indices were obtained under a fixed average glandular dose (AGD) and a fixed image contrast. Our results indicated that when the image contrast and AGD was constant, for phantom thinner than 5 cm, an increase of the tube voltage did not improve the noise property of images very much. The results also showed that image property with the target/filter Mo/Rh was better than that with Mo/Mo for phantom thicker than 4 cm. In general, it is said that high tube voltage delivers improved noise property. Our result indicates that this common theory is not realized with the x-ray energy level for mammography.

  16. Digital Mammography: Improvements in Breast Cancer Diagnostic

    NASA Astrophysics Data System (ADS)

    Montaño Zetina, Luis Manuel

    2006-01-01

    X-ray mammography is the most sensitive imaging technique for early detection of breast cancer (diagnostics). It is performed by a radiological system equipped with a rotating molybdenum (Mo) anode tube with an additional Mo filter. In the production of X-ray, bremsstrahlung photons produce an intense diffuse radiation, affecting the contrast between normal and cancerous tissue. So it is known that a good mammographic imaging can help to detect cancer in the first stages avoiding surgery, amputation or even death. In the last years there has been some developments in new imaging techniques to improve the contrast spatial resolution between different tissues: digital imaging, or the so call digital mammography. Digital mammographic imaging is considered an improvement in the prevention of breast cancer due to the advantages it offers.

  17. Quality indicators of image-based overlay

    NASA Astrophysics Data System (ADS)

    Chen, Yen-Liang; Huang, Jacky; Lee, Rita; Wang, Chen-Ming; Ke, Chih-Ming; Gau, Tsai-Sheng

    2012-03-01

    A new method for indicating the image quality of overlay measurement is proposed in this paper. Due to the constraint of the overlay control tolerance, the overlay metrology requirement has become very stringent. Current indicators such as the total measurement uncertainty (TMU) are insufficient to guarantee a good overlay measurement. This paper describes two quality indicators, the contrast index (CI) and the asymmetry index (AI). The CI is a crucial quality indicator that affects the overlay accuracy greatly. The AI, based on an imaging process with modified cross-correlation operation, shows alignment mark robustness in both the x and the y directions. For determination of the best recipe, the box-in-box overlay marks are measured to obtain the images with different conditions. The conventional TMU indicators are used first to sieve out the better choices. Then the CI and AI can help to judge whether the overlay results are reliable and can be applied to monitoring of process variations.

  18. Lessions learned in WISE image quality

    NASA Astrophysics Data System (ADS)

    Kendall, Martha; Duval, Valerie G.; Larsen, Mark F.; Heinrichsen, Ingolf H.; Esplin, Roy W.; Shannon, Mark; Wright, Edward L.

    2010-08-01

    The Wide-Field Infrared Survey Explorer (WISE) mission launched in December of 2009 is a true success story. The mission is performing beyond expectations on-orbit and maintained cost and schedule throughout. How does such a thing happen? A team constantly focused on mission success is a key factor. Mission success is more than a program meeting its ultimate science goals; it is also meeting schedule and cost goals to avoid cancellation. The WISE program can attribute some of its success in achieving the image quality needed to meet science goals to lessons learned along the way. A requirement was missed in early decomposition, the absence of which would have adversely affected end-to-end system image quality. Fortunately, the ability of the cross-organizational team to focus on fixing the problem without pointing fingers or waiting for paperwork was crucial in achieving a timely solution. Asking layman questions early in the program could have revealed requirement flowdown misunderstandings between spacecraft control stability and image processing needs. Such is the lesson learned with the WISE spacecraft Attitude Determination & Control Subsystem (ADCS) jitter control and the image data reductions needs. Spacecraft motion can affect image quality in numerous ways. Something as seemingly benign as different terminology being used by teammates in separate groups working on data reduction, spacecraft ADCS, the instrument, mission operations, and the science proved to be a risk to system image quality. While the spacecraft was meeting the allocated jitter requirement , the drift rate variation need was not being met. This missing need was noticed about a year before launch and with a dedicated team effort, an adjustment was made to the spacecraft ADCS control. WISE is meeting all image quality requirements on-orbit thanks to a diligent team noticing something was missing before it was too late and applying their best effort to find a solution.

  19. Image Quality Assessment using web services

    NASA Astrophysics Data System (ADS)

    Malapert, J.-C.; Magnard, F.

    2006-07-01

    Nowadays, image processing can be very demanding in terms of disk space, CPU, memory, resources which laboratory might not necessarily have. As a result, TERAPIX (Institut d'Astrophysique de Paris) has developed an image quality assessment tool which can be accessed by a web service. This service is developed with Virtual Observatory standards in mind, and will be accessible through standard protocols in order to ensure its interoperability with other software.

  20. Robust approach for color image quality assessment

    NASA Astrophysics Data System (ADS)

    Le Callet, Patrick; Barba, Dominique

    2003-06-01

    This paper presents a visual color image quality metric assessment with full reference image. The metric is highly based on human visual system properties in order to get the best correspondence with human judgements. Contrary to some others objective criteria, it doesn't use any a priori knowledge on the type of introduced degradations. So the main interest of the metric is on its ability to produce robust results independently of the distortions. The metric can be decomposed into two steps. The first one projects each images, the reference one and the distorted one, in a perceptual space. The second step achieves the pooling of errors between perceptual representation of two images in order to get a score for the overall quality. Since we have shown that these two steps have equivalent importance regarding metric performance, we have particularly paid attention in correct balancing when designing the two steps. Especially, for the second one, that is generally limited to poor consideration in literature, we have developed some new original approaches . We compare results of the metric with human judgments on images distorted with different compression schemes. High performances are obtained leading to assure that the metric is robust, so this approach constitutes an alternative useful tool to PSNR for image quality assessment.

  1. Subjective matters: from image quality to image psychology

    NASA Astrophysics Data System (ADS)

    Fedorovskaya, Elena A.; De Ridder, Huib

    2013-03-01

    From the advent of digital imaging through several decades of studies, the human vision research community systematically focused on perceived image quality and digital artifacts due to resolution, compression, gamma, dynamic range, capture and reproduction noise, blur, etc., to help overcome existing technological challenges and shortcomings. Technological advances made digital images and digital multimedia nearly flawless in quality, and ubiquitous and pervasive in usage, provide us with the exciting but at the same time demanding possibility to turn to the domain of human experience including higher psychological functions, such as cognition, emotion, awareness, social interaction, consciousness and Self. In this paper we will outline the evolution of human centered multidisciplinary studies related to imaging and propose steps and potential foci of future research.

  2. Image Fusion and Image Quality Assessment of Fused Images

    NASA Astrophysics Data System (ADS)

    Han, Z.; Tang, X.; Gao, X.; Hu, F.

    2013-07-01

    It is of great value to fuse a high-resolution panchromatic image and low-resolution multi-spectral images for object recognition. In the paper, tow frames of remotely sensed imagery, including ZY03 and SPOT05, are selected as the source data. Four fusion methods, including Brovey, PCA, Pansharp, and SFIM, are used to fuse the images of multispectral bands and panchromatic band. Three quantitative indicators were calculated and analyzed, that is, gradient, correlation coefficient and deviation. According to comprehensive evaluation and comparison, the best effect is SFIM transformation, combined with fusion image through four transformation methods.

  3. FFDM image quality assessment using computerized image texture analysis

    NASA Astrophysics Data System (ADS)

    Berger, Rachelle; Carton, Ann-Katherine; Maidment, Andrew D. A.; Kontos, Despina

    2010-04-01

    Quantitative measures of image quality (IQ) are routinely obtained during the evaluation of imaging systems. These measures, however, do not necessarily correlate with the IQ of the actual clinical images, which can also be affected by factors such as patient positioning. No quantitative method currently exists to evaluate clinical IQ. Therefore, we investigated the potential of using computerized image texture analysis to quantitatively assess IQ. Our hypothesis is that image texture features can be used to assess IQ as a measure of the image signal-to-noise ratio (SNR). To test feasibility, the "Rachel" anthropomorphic breast phantom (Model 169, Gammex RMI) was imaged with a Senographe 2000D FFDM system (GE Healthcare) using 220 unique exposure settings (target/filter, kVs, and mAs combinations). The mAs were varied from 10%-300% of that required for an average glandular dose (AGD) of 1.8 mGy. A 2.5cm2 retroareolar region of interest (ROI) was segmented from each image. The SNR was computed from the ROIs segmented from images linear with dose (i.e., raw images) after flat-field and off-set correction. Image texture features of skewness, coarseness, contrast, energy, homogeneity, and fractal dimension were computed from the Premium ViewTM postprocessed image ROIs. Multiple linear regression demonstrated a strong association between the computed image texture features and SNR (R2=0.92, p<=0.001). When including kV, target and filter as additional predictor variables, a stronger association with SNR was observed (R2=0.95, p<=0.001). The strong associations indicate that computerized image texture analysis can be used to measure image SNR and potentially aid in automating IQ assessment as a component of the clinical workflow. Further work is underway to validate our findings in larger clinical datasets.

  4. Geometric assessment of image quality using digital image registration techniques

    NASA Technical Reports Server (NTRS)

    Tisdale, G. E.

    1976-01-01

    Image registration techniques were developed to perform a geometric quality assessment of multispectral and multitemporal image pairs. Based upon LANDSAT tapes, accuracies to a small fraction of a pixel were demonstrated. Because it is insensitive to the choice of registration areas, the technique is well suited to performance in an automatic system. It may be implemented at megapixel-per-second rates using a commercial minicomputer in combination with a special purpose digital preprocessor.

  5. Method for inserting noise in digital mammography to simulate reduction in radiation dose

    NASA Astrophysics Data System (ADS)

    Borges, Lucas R.; de Oliveira, Helder C. R.; Nunes, Polyana F.; Vieira, Marcelo A. C.

    2015-03-01

    The quality of clinical x-ray images is closely related to the radiation dose used in the imaging study. The general principle for selecting the radiation is ALARA ("as low as reasonably achievable"). The practical optimization, however, remains challenging. It is well known that reducing the radiation dose increases the quantum noise, which could compromise the image quality. In order to conduct studies about dose reduction in mammography, it would be necessary to acquire repeated clinical images, from the same patient, with different dose levels. However, such practice would be unethical due to radiation related risks. One solution is to simulate the effects of dose reduction in clinical images. This work proposes a new method, based on the Anscombe transformation, which simulates dose reduction in digital mammography by inserting quantum noise into clinical mammograms acquired with the standard radiation dose. Thus, it is possible to simulate different levels of radiation doses without exposing the patient to new levels of radiation. Results showed that the achieved quality of simulated images generated with our method is the same as when using other methods found in the literature, with the novelty of using the Anscombe transformation for converting signal-independent Gaussian noise into signal-dependent quantum noise.

  6. Image Quality in Analog and Digital Microtechniques.

    ERIC Educational Resources Information Center

    White, William

    1991-01-01

    Discusses the basic principles of the application of microfilm (analog) and electronic (digital) technologies for data storage. Image quality is examined, searching and retrieval capabilities are considered, and hardcopy output resolution is described. It is concluded that microfilm is still the preferred archival medium. (5 references) (LRW)

  7. Image Quality Indicator for Infrared Inspections

    NASA Technical Reports Server (NTRS)

    Burke, Eric

    2011-01-01

    The quality of images generated during an infrared thermal inspection depends on many system variables, settings, and parameters to include the focal length setting of the IR camera lens. If any relevant parameter is incorrect or sub-optimal, the resulting IR images will usually exhibit inherent unsharpness and lack of resolution. Traditional reference standards and image quality indicators (IQIs) are made of representative hardware samples and contain representative flaws of concern. These standards are used to verify that representative flaws can be detected with the current IR system settings. However, these traditional standards do not enable the operator to quantify the quality limitations of the resulting images, i.e. determine the inherent maximum image sensitivity and image resolution. As a result, the operator does not have the ability to optimize the IR inspection system prior to data acquisition. The innovative IQI described here eliminates this limitation and enables the operator to objectively quantify and optimize the relevant variables of the IR inspection system, resulting in enhanced image quality with consistency and repeatability in the inspection application. The IR IQI consists of various copper foil features of known sizes that are printed on a dielectric non-conductive board. The significant difference in thermal conductivity between the two materials ensures that each appears with a distinct grayscale or brightness in the resulting IR image. Therefore, the IR image of the IQI exhibits high contrast between the copper features and the underlying dielectric board, which is required to detect the edges of the various copper features. The copper features consist of individual elements of various shapes and sizes, or of element-pairs of known shapes and sizes and with known spacing between the elements creating the pair. For example, filled copper circles with various diameters can be used as individual elements to quantify the image sensitivity limit. Copper line-pairs of various sizes where the line width is equivalent to the spacing between the lines can be used as element-pairs to quantify the image resolution limit.

  8. Scene reduction for subjective image quality assessment

    NASA Astrophysics Data System (ADS)

    Lewandowska (Tomaszewska), Anna

    2016-01-01

    Evaluation of image quality is important for many image processing systems, such as those used for acquisition, compression, restoration, enhancement, or reproduction. Its measurement is often accompanied by user studies, in which a group of observers rank or rate results of several algorithms. Such user studies, known as subjective image quality assessment experiments, can be very time consuming and do not guarantee conclusive results. This paper is intended to help design an efficient and rigorous quality assessment experiment. We propose a method of limiting the number of scenes that need to be tested, which can significantly reduce the experimental effort and still capture relevant scene-dependent effects. To achieve it, we employ a clustering technique and evaluate it on the basis of compactness and separation criteria. The correlation between the results obtained from a set of images in an initial database and the results received from reduced experiment are analyzed. Finally, we propose a procedure for reducing the initial scenes number. Four different assessment techniques were tested: single stimulus, double stimulus, forced choice, and similarity judgments. We conclude that in most cases, 9 to 12 judgments per evaluated algorithm for a large scene collection is sufficient to reduce the initial set of images.

  9. Characterization of scatter in digital mammography from physical measurements

    SciTech Connect

    Leon, Stephanie M. Wagner, Louis K.; Brateman, Libby F.

    2014-06-15

    Purpose: That scattered radiation negatively impacts the quality of medical radiographic imaging is well known. In mammography, even slight amounts of scatter reduce the high contrast required for subtle soft-tissue imaging. In current clinical mammography, image contrast is partially improved by use of an antiscatter grid. This form of scatter rejection comes with a sizeable dose penalty related to the concomitant elimination of valuable primary radiation. Digital mammography allows the use of image processing as a method of scatter correction that might avoid effects that negatively impact primary radiation, while potentially providing more contrast improvement than is currently possible with a grid. For this approach to be feasible, a detailed characterization of the scatter is needed. Previous research has modeled scatter as a constant background that serves as a DC bias across the imaging surface. The goal of this study was to provide a more substantive data set for characterizing the spatially-variant features of scatter radiation at the image detector of modern mammography units. Methods: This data set was acquired from a model of the radiation beam as a matrix of very narrow rays or pencil beams. As each pencil beam penetrates tissue, the pencil widens in a predictable manner due to the production of scatter. The resultant spreading of the pencil beam at the detector surface can be characterized by two parameters: mean radial extent (MRE) and scatter fraction (SF). The SF and MRE were calculated from measurements obtained using the beam stop method. Two digital mammography units were utilized, and the SF and MRE were found as functions of target, filter, tube potential, phantom thickness, and presence or absence of a grid. These values were then used to generate general equations allowing the SF and MRE to be calculated for any combination of the above parameters. Results: With a grid, the SF ranged from a minimum of about 0.05 to a maximum of about 0.16, and the MRE ranged from about 3 to 13 mm. Without a grid, the SF ranged from a minimum of 0.25 to a maximum of 0.52, and the MRE ranged from about 20 to 45 mm. The SF with a grid demonstrated a mild dependence on target/filter combination and kV, whereas the SF without a grid was independent of these factors. The MRE demonstrated a complex relationship as a function of kV, with notable difference among target/filter combinations. The primary source of change in both the SF and MRE was phantom thickness. Conclusions: Because breast tissue varies spatially in physical density and elemental content, the effective thickness of breast tissue varies spatially across the imaging field, resulting in a spatially-variant scatter distribution in the imaging field. The data generated in this study can be used to characterize the scatter contribution on a point-by-point basis, for a variety of different techniques.

  10. Modeling and simulation of Positron Emission Mammography (PEM) based on double-sided CdTe strip detectors

    NASA Astrophysics Data System (ADS)

    Ozsahin, I.; Unlu, M. Z.

    2014-03-01

    Breast cancer is the most common leading cause of cancer death among women. Positron Emission Tomography (PET) Mammography, also known as Positron Emission Mammography (PEM), is a method for imaging primary breast cancer. Over the past few years, PEMs based on scintillation crystals dramatically increased their importance in diagnosis and treatment of early stage breast cancer. However, these detectors have significant limitations like poor energy resolution resulting with false-negative result (missed cancer), and false-positive result which leads to suspecting cancer and suggests an unnecessary biopsy. In this work, a PEM scanner based on CdTe strip detectors is simulated via the Monte Carlo method and evaluated in terms of its spatial resolution, sensitivity, and image quality. The spatial resolution is found to be ~ 1 mm in all three directions. The results also show that CdTe strip detectors based PEM scanner can produce high resolution images for early diagnosis of breast cancer.

  11. Prediction of Viking lander camera image quality

    NASA Technical Reports Server (NTRS)

    Huck, F. O.; Burcher, E. E.; Jobson, D. J.; Wall, S. D.

    1976-01-01

    Formulations are presented that permit prediction of image quality as a function of camera performance, surface radiance properties, and lighting and viewing geometry. Predictions made for a wide range of surface radiance properties reveal that image quality depends strongly on proper camera dynamic range command and on favorable lighting and viewing geometry. Proper camera dynamic range commands depend mostly on the surface albedo that will be encountered. Favorable lighting and viewing geometries depend mostly on lander orientation with respect to the diurnal sun path over the landing site, and tend to be independent of surface albedo and illumination scattering function. Side lighting with low sun elevation angles (10 to 30 deg) is generally favorable for imaging spatial details and slopes, whereas high sun elevation angles are favorable for measuring spectral reflectances.

  12. Naturalness and interestingness of test images for visual quality evaluation

    NASA Astrophysics Data System (ADS)

    Halonen, Raisa; Westman, Stina; Oittinen, Pirkko

    2011-01-01

    Balanced and representative test images are needed to study perceived visual quality in various application domains. This study investigates naturalness and interestingness as image quality attributes in the context of test images. Taking a top-down approach we aim to find the dimensions which constitute naturalness and interestingness in test images and the relationship between these high-level quality attributes. We compare existing collections of test images (e.g. Sony sRGB images, ISO 12640 images, Kodak images, Nokia images and test images developed within our group) in an experiment combining quality sorting and structured interviews. Based on the data gathered we analyze the viewer-supplied criteria for naturalness and interestingness across image types, quality levels and judges. This study advances our understanding of subjective image quality criteria and enables the validation of current test images, furthering their development.

  13. Quantitative Statistical Methods for Image Quality Assessment

    PubMed Central

    Dutta, Joyita; Ahn, Sangtae; Li, Quanzheng

    2013-01-01

    Quantitative measures of image quality and reliability are critical for both qualitative interpretation and quantitative analysis of medical images. While, in theory, it is possible to analyze reconstructed images by means of Monte Carlo simulations using a large number of noise realizations, the associated computational burden makes this approach impractical. Additionally, this approach is less meaningful in clinical scenarios, where multiple noise realizations are generally unavailable. The practical alternative is to compute closed-form analytical expressions for image quality measures. The objective of this paper is to review statistical analysis techniques that enable us to compute two key metrics: resolution (determined from the local impulse response) and covariance. The underlying methods include fixed-point approaches, which compute these metrics at a fixed point (the unique and stable solution) independent of the iterative algorithm employed, and iteration-based approaches, which yield results that are dependent on the algorithm, initialization, and number of iterations. We also explore extensions of some of these methods to a range of special contexts, including dynamic and motion-compensated image reconstruction. While most of the discussed techniques were developed for emission tomography, the general methods are extensible to other imaging modalities as well. In addition to enabling image characterization, these analysis techniques allow us to control and enhance imaging system performance. We review practical applications where performance improvement is achieved by applying these ideas to the contexts of both hardware (optimizing scanner design) and image reconstruction (designing regularization functions that produce uniform resolution or maximize task-specific figures of merit). PMID:24312148

  14. Medical Imaging Image Quality Assessment with Monte Carlo Methods

    NASA Astrophysics Data System (ADS)

    Michail, C. M.; Karpetas, G. E.; Fountos, G. P.; Kalyvas, N. I.; Martini, Niki; Koukou, Vaia; Valais, I. G.; Kandarakis, I. S.

    2015-09-01

    The aim of the present study was to assess image quality of PET scanners through a thin layer chromatography (TLC) plane source. The source was simulated using a previously validated Monte Carlo model. The model was developed by using the GATE MC package and reconstructed images obtained with the STIR software for tomographic image reconstruction, with cluster computing. The PET scanner simulated in this study was the GE DiscoveryST. A plane source consisted of a TLC plate, was simulated by a layer of silica gel on aluminum (Al) foil substrates, immersed in 18F-FDG bath solution (1MBq). Image quality was assessed in terms of the Modulation Transfer Function (MTF). MTF curves were estimated from transverse reconstructed images of the plane source. Images were reconstructed by the maximum likelihood estimation (MLE)-OSMAPOSL algorithm. OSMAPOSL reconstruction was assessed by using various subsets (3 to 21) and iterations (1 to 20), as well as by using various beta (hyper) parameter values. MTF values were found to increase up to the 12th iteration whereas remain almost constant thereafter. MTF improves by using lower beta values. The simulated PET evaluation method based on the TLC plane source can be also useful in research for the further development of PET and SPECT scanners though GATE simulations.

  15. Dried fruits quality assessment by hyperspectral imaging

    NASA Astrophysics Data System (ADS)

    Serranti, Silvia; Gargiulo, Aldo; Bonifazi, Giuseppe

    2012-05-01

    Dried fruits products present different market values according to their quality. Such a quality is usually quantified in terms of freshness of the products, as well as presence of contaminants (pieces of shell, husk, and small stones), defects, mould and decays. The combination of these parameters, in terms of relative presence, represent a fundamental set of attributes conditioning dried fruits humans-senses-detectable-attributes (visual appearance, organolectic properties, etc.) and their overall quality in terms of marketable products. Sorting-selection strategies exist but sometimes they fail when a higher degree of detection is required especially if addressed to discriminate between dried fruits of relatively small dimensions and when aiming to perform an "early detection" of pathogen agents responsible of future moulds and decays development. Surface characteristics of dried fruits can be investigated by hyperspectral imaging (HSI). In this paper, specific and "ad hoc" applications addressed to propose quality detection logics, adopting a hyperspectral imaging (HSI) based approach, are described, compared and critically evaluated. Reflectance spectra of selected dried fruits (hazelnuts) of different quality and characterized by the presence of different contaminants and defects have been acquired by a laboratory device equipped with two HSI systems working in two different spectral ranges: visible-near infrared field (400-1000 nm) and near infrared field (1000-1700 nm). The spectra have been processed and results evaluated adopting both a simple and fast wavelength band ratio approach and a more sophisticated classification logic based on principal component (PCA) analysis.

  16. Image quality control in breast ultrasound.

    PubMed

    Lagalla, R; Midiri, M

    1998-05-01

    Sonography is well suited for breast studies. Adequate equipment is needed to acquire high quality images because several technical factors influence ultrasound images. Thus, the use of high frequency dynamic scanning probes, the ultrasound beam focusing corrected for the near field, the adjustment of the gain and image contrast may all interfere with ultrasound beam reflection and scattering, determined by the heterogeneity of the gland parenchyma. In the last few years, a line of ultrasound equipment dedicated to this kind of application has been developed with 'small parts' transducers and frequencies ranging 10-13 MHz. These units can improve the evaluation of superficial structures and provide diagnostic results that conventional equipment cannot achieve. The higher the quality, the more a sonographic image corresponds to real anatomy. This capability depends on the different kinds of system resolution. Axial spatial resolution is the capability to resolve discrete structures along the beam axis. Pulse length is inversely proportional to frequency and thus, the higher the transducer frequency, the better the axial resolution. However, the increase in frequency reduces the depth of penetration of the ultrasound beam. The spectrum of frequencies emitted by the crystal has been recently modified in order to obtain a good trade-off between the beam resolution and its penetration. Indeed, the development of the multifrequency technology allowed to improve the near field resolution while retaining a good penetration into the distant field. Furthermore, the use of compound ceramics with a broad bandwidth helps Doppler analysis because flow studies are optimized by low frequencies, whereas two-dimensional morphologic imaging is optimized by high frequencies. Lateral spatial resolution is the capability to resolve discrete structures perpendicular or lateral to the beam axis. This parameter strictly depends on the size of the ultrasound beam section and it is optimal only in the focal area. Therefore, it improves with narrow beams. Several transducers are available in breast sonography, but the most adequate one is currently the annular transducer. The equipment should be able to detect even slight differences in acoustic impedance between the several breast tissues. This may be obtained by optimizing the dynamic range and the pre- and postprocessing setting. Apart from equipment, two other technical factors should be optimized to obtain high quality images, namely beam intensity and gain curve. A new Doppler technique has been recently introduced: power Doppler, which allows the demonstration of breast nodule vascularization with higher sensitivity than color Doppler. Finally, a rigorous examination technique is required to obtain high quality images. In the last few years, several quality assurance programs have been introduced. Dedicated phantoms are generally used. Recently, computer systems have been also developed. PMID:9652527

  17. Beyond the mammography debate: a moderate perspective

    PubMed Central

    Kaniklidis, C

    2015-01-01

    After some decades of contention, one can almost despair and conclude that (paraphrasing) “the mammography debate you will have with you always.” Against that sentiment, in this review I argue, after reflecting on some of the major themes of this long-standing debate, that we must begin to move beyond the narrow borders of claim and counterclaim to seek consensus on what the balance of methodologically sound and critically appraised evidence demonstrates, and also to find overlooked underlying convergences; after acknowledging the reality of some residual and non-trivial harms from mammography, to promote effective strategies for harm mitigation; and to encourage deployment of new screening modalities that will render many of the issues and concerns in the debate obsolete. To these ends, I provide a sketch of what this looking forward and beyond the current debate might look like, leveraging advantages from abbreviated breast magnetic resonance imaging technologies (such as the ultrafast and twist protocols) and from digital breast tomosynthesis—also known as three-dimensional mammography. I also locate the debate within the broader context of mammography in the real world as it plays out not for the disputants, but for the stakeholders themselves: the screening-eligible patients and the physicians in the front lines who are charged with enabling both the acts of screening and the facts of screening at their maximally objective and patient-accessible levels to facilitate informed decisions. PMID:26089721

  18. Clinical mammography at the SYRMEP beam line

    NASA Astrophysics Data System (ADS)

    Castelli, E.; Arfelli, F.; Dreossi, D.; Longo, R.; Rokvic, T.; Cova, M. A.; Quaia, E.; Tonutti, M.; Zanconati, F.; Abrami, A.; Chenda, V.; Menk, R. H.; Quai, E.; Tromba, G.; Bregant, P.; de Guarrini, F.

    2007-03-01

    The synchrotron radiation (SR) clinical mammography, using as detector a commercial screen-film system, is a further crucial step of the SYRMEP (Synchrotron Radiation for Medical Physics) Project. In March and April of this year, mammographic examinations on nine patients have been carried out with X-rays from ELETTRA, the SR laboratory in Trieste, Italy. The facility for Phase Contrast (PhC) SR mammography is now operative in patient mode and is used for patient examinations, producing breast images in different projections. The entire procedure is automated and is able to achieve the correct exposure on the film taking into consideration the requested limits on the applied dose, as a function of the thickness and glandularity of the breast. The SR mammography shows a higher spatial resolution and contrast detail visibility, in comparison with the conventional analog or digital mammography, with a comparable or lower dose. In some cases of these first nine patients, the radiologists have clarified the ambiguity of the previous examinations. These preliminary results are encouraging and a complete evaluation of the clinical impact of the new method is in progress.

  19. Geographic Access and the Use of Screening Mammography

    PubMed Central

    Elkin, Elena B.; Ishill, Nicole M.; Snow, Jacqueline G.; Panageas, Katherine S.; Bach, Peter B.; Liberman, Laura; Wang, Fahui; Schrag, Deborah

    2011-01-01

    Background Screening mammography rates vary geographically and have recently declined. Inadequate mammography resources in some areas may impair access to this technology. We assessed the relationship between availability of mammography machines and the use of screening. Methods The location and number of all mammography machines in the US were identified from US Food and Drug Administration records of certified facilities. Inadequate capacity was defined as <1.2 mammography machines per 10,000 women aged 40 or older, the threshold required to meet the Healthy People 2010 target screening rate. The impact of capacity on utilization was evaluated in two cohorts: female respondents age 40 or older to the 2006 Behavioral Risk Factor Surveillance System survey (BRFSS) and a 5% nationwide sample of female Medicare beneficiaries age 65 or older in 2004–2005. Results About 9% of women in the BRFSS cohort and 13% of women in the Medicare cohort lived in counties with <1.2 mammography machines per 10,000 women age 40 or older. In both cohorts, residence in a county with inadequate mammography capacity was associated with lower odds of a recent mammogram (adjusted odds ratio [AOR] in BRFSS: 0.89, 95% CI 0.80 – 0.98, p<0.05; AOR in Medicare: 0.86, 95% CI 0.85 – 0.87, p<0.05), controlling for demographic and health care characteristics. Conclusion In counties with few or no mammography machines, limited availability of imaging resources may be a barrier to screening. Efforts to increase the number of machines in low-capacity areas may improve mammography rates and reduce geographic disparities in breast cancer screening. PMID:20195174

  20. [Evaluation of the 1Shot Phantom dedicated to the mammography system using FCR].

    PubMed

    Nagashima, Chieko; Uchiyama, Nachiko; Moriyama, Noriyuki; Nagata, Mio; Kobayashi, Hiroyuki; Sankoda, Katsuhiro; Saotome, Shigeru; Tagi, Masahiro; Kusunoki, Tetsurou

    2009-07-20

    Currently daily quality control (QC) tests for mammography systems are generally evaluated by using visual analysis phantoms, which of course means subjective measurement. In our study, however, we evaluated a novel digital phantom, the 1Shot Phantom M plus (1Shot Phantom), together with automatic analysis software dedicated for mammography systems using Fuji computed radiography (FCR). The digital phantom enables objective evaluation by providing for actual physical measurement rather than subjective visual assessment. We measured 1) contrast to noise ratio (CNR), 2) image receptor homogeneity, 3) missed tissue at chest wall side, 4) modulation transfer function (MTF), and 5) geometric distortion utilizing the 1Shot Phantom. We then compared the values obtained using the 1Shot Phantom with values obtained from the European guidelines and International Electrotechnical Commission (IEC) standards. In addition, we evaluated the convenience of using the digital phantom. The values utilizing the 1Shot Phantom and those from the European guidelines and IEC standards were consistent, but the QC tests for the European guidelines and IEC standards methods took about six hours while the same QC tests using the 1Shot Phantom took 10 minutes or less including exposure of the phantom image, measurement, and analysis. In conclusion, the digital phantom and dedicated software proved very useful and produced improved analysis for mammography systems using FCR in clinical daily QC testing because of their objectivity and substantial time-saving convenience. PMID:19661726

  1. BREAST: a novel method to improve the diagnostic efficacy of mammography

    NASA Astrophysics Data System (ADS)

    Brennan, P. C.; Tapia, K.; Ryan, J.; Lee, W.

    2013-03-01

    High quality breast imaging and accurate image assessment are critical to the early diagnoses, treatment and management of women with breast cancer. Breast Screen Reader Assessment Strategy (BREAST) provides a platform, accessible by researchers and clinicians world-wide, which will contain image data bases, algorithms to assess reader performance and on-line systems for image evaluation. The platform will contribute to the diagnostic efficacy of breast imaging in Australia and beyond on two fronts: reducing errors in mammography, and transforming our assessment of novel technologies and techniques. Mammography is the primary diagnostic tool for detecting breast cancer with over 800,000 women X-rayed each year in Australia, however, it fails to detect 30% of breast cancers with a number of missed cancers being visible on the image [1-6]. BREAST will monitor the mistakes, identify reasons for mammographic errors, and facilitate innovative solutions to reduce error rates. The BREAST platform has the potential to enable expert assessment of breast imaging innovations, anywhere in the world where experts or innovations are located. Currently, innovations are often being assessed by limited numbers of individuals who happen to be geographically located close to the innovation, resulting in equivocal studies with low statistical power. BREAST will transform this current paradigm by enabling large numbers of experts to assess any new method or technology using our embedded evaluation methods. We are confident that this world-first system will play an important part in the future efficacy of breast imaging.

  2. Image quality assessment and human visual system

    NASA Astrophysics Data System (ADS)

    Gao, Xinbo; Lu, Wen; Tao, Dacheng; Li, Xuelong

    2010-07-01

    This paper summaries the state-of-the-art of image quality assessment (IQA) and human visual system (HVS). IQA provides an objective index or real value to measure the quality of the specified image. Since human beings are the ultimate receivers of visual information in practical applications, the most reliable IQA is to build a computational model to mimic the HVS. According to the properties and cognitive mechanism of the HVS, the available HVS-based IQA methods can be divided into two categories, i.e., bionics methods and engineering methods. This paper briefly introduces the basic theories and development histories of the above two kinds of HVS-based IQA methods. Finally, some promising research issues are pointed out in the end of the paper.

  3. Influence of the characteristic curve on the clinical image quality and patient absorbed dose in lumbar spine radiography

    NASA Astrophysics Data System (ADS)

    Tingberg, Anders; Herrmann, Clemens; Lanhede, Birgitta; Almen, Anja; Mattsson, Saron; Panzer, Werner; Besjakov, Jack; Mansson, Lars G.; Kheddache, Susanne; Zankl, Maria

    2001-06-01

    The 'European Guidelines on Quality Criteria for Diagnostic Radiographic Images' do not address the choice of film characteristic (H/D) curve, which is an important parameter for the description of a radiographic screen-film system. Since it is not possible to investigate this influence by taking repeated exposures of the same patients on films with systematically varied H/D curves, patient images of lumbar spine were digitised in the current study. The image contrast was altered by digital image processing techniques, simulating images with H/D curves varying from flat over standard latitude to a film type steeper than a mammography film. The manipulated images were printed on film for evaluation. Seven European radiologists evaluated the clinical image quality of in total 224 images by analysing the fulfilment of the European Image Criteria and by visual grading analysis of the images. The results show that the local quality can be significantly improved by the application of films with a steeper film H/D curve compared to the standard latitude film. For images with an average optical density of about 1.25, the application of the steeper film results in a reduction of patient absorbed dose by about 10-15% without a loss of diagnostically relevant image information. The results also show that the patient absorbed dose reduction obtained by altering the tube voltage from 70 kV to 90 kV coincides with a loss of image information that cannot be compensated for by simply changing the shape of the H/D curve.

  4. Visual discrimination model for digital mammography

    NASA Astrophysics Data System (ADS)

    Johnson, Jeffrey P.; Lubin, Jeffrey; Krupinski, Elizabeth A.; Peterson, Heidi A.; Roehrig, Hans; Baysinger, Andrew

    1999-05-01

    Numerous studies have been conducted to determine experimentally the effects of image processing and display parameters on the diagnostic performance of radiologists. Comprehensive optimization of imaging systems for digital mammography based solely on measurements of reader performance is impractical, however, due to the large number of interdependent variables to be tested. A reliable, efficient alternative is needed to improve the evaluation and optimization of new imaging technologies. The Sarnoff JNDmetrixTM Visual Discrimination Model (VDM) is a computational, just-noticeable difference model of human vision that has been applied successfully to predict performance in various nonmedical detection and rating tasks. To test the applicability of the VDM to specific detection tasks in digital mammography, two observer performance studies were conducted. In the first study, effects of display tone scale and peak luminance on the detectability of microcalcifications were evaluated. The VDM successfully predicted improvements in reader performance for perceptually linearized tone scales and higher display luminances. In the second study, the detectability of JPEG and wavelet compression artifacts was evaluated, and performance ratings were again found to be highly correlated with VDM predictions. These results suggest that the VDM would be useful in the assessment and optimization of new imaging and compression technologies for digital mammography.

  5. Computer-Aided Diagnosis in Mammography Using Content-based Image Retrieval Approaches: Current Status and Future Perspectives

    PubMed Central

    Zheng, Bin

    2010-01-01

    As the rapid advance of digital imaging technologies, the content-based image retrieval (CBIR) has became one of the most vivid research areas in computer vision. In the last several years, developing computer-aided detection and/or diagnosis (CAD) schemes that use CBIR to search for the clinically relevant and visually similar medical images (or regions) depicting suspicious lesions has also been attracting research interest. CBIR-based CAD schemes have potential to provide radiologists with “visual aid” and increase their confidence in accepting CAD-cued results in the decision making. The CAD performance and reliability depends on a number of factors including the optimization of lesion segmentation, feature selection, reference database size, computational efficiency, and relationship between the clinical relevance and visual similarity of the CAD results. By presenting and comparing a number of approaches commonly used in previous studies, this article identifies and discusses the optimal approaches in developing CBIR-based CAD schemes and assessing their performance. Although preliminary studies have suggested that using CBIR-based CAD schemes might improve radiologists’ performance and/or increase their confidence in the decision making, this technology is still in the early development stage. Much research work is needed before the CBIR-based CAD schemes can be accepted in the clinical practice. PMID:20305801

  6. Multiple-reader studies, digital mammography, computer-aided diagnosis, and the Holy Grail of imaging physics: II

    NASA Astrophysics Data System (ADS)

    Beiden, Sergey V.; Wagner, Robert F.; Campbell, Gregory; Metz, Charles E.; Jiang, Yulei; Chan, Heang-Ping

    2001-06-01

    The metaphor of the Holy Grail is used here to refer to the classic and elusive problem in medical imaging of predicting the ranking of the clinical performance of competing imaging modalities from the ranking obtained from physical laboratory measurements and signal-detection analysis, or from simple phantom studies. We show how the use of the multiple-reader, multiple-case (MRMC) ROC paradigm and new analytical techniques allows this masking effect to be quantified in terms of components-of-variance models. Moreover, we demonstrate how the components of variance associated with reader variability may be reduced when readers have the benefit of computer-assist reading aids. The remaining variability will be due to the case components, and these reflect the contribution of the technology without the masking effect of the reader. This suggests that prediction of clinical ranking of imaging systems in terms of physical measurements may become a much more tractable task in a world that includes MRMC ROC analysis of performance of radiologists with the advantage of computer-assisted reading.

  7. A glass-ceramic plate for mammography.

    SciTech Connect

    Johnson, J. A.; Schweizer, S.; Lubinsky, A. R.; Nuclear Engineering Division; Univ. of Paderborn; State Univ. of New York at Stony Brook

    2007-01-01

    We developed translucent glass-ceramic image plates for digital mammography. The glass ceramics are based on europium-doped fluorozirconate glasses, which were additionally doped with chlorine to initiate the nucleation of barium chloride nanoparticles therein. The X-ray image is stored in the form of stable electron-hole pairs, which can be read out afterwards with a scanning laser beam in a 'photostimulated luminescence' (PSL) process. Measurements of the required stimulating exposure, integrated PSL signal, and optical light spreading of the stimulating laser light were performed to allow projection of the detective quantum efficiency (DQE) for the proposed X-ray storage phosphor system. The projected DQE is compared with commercially available electronic mammography systems.

  8. Toward Clinically Compatible Phase-Contrast Mammography

    PubMed Central

    Scherer, Kai; Willer, Konstantin; Gromann, Lukas; Birnbacher, Lorenz; Braig, Eva; Grandl, Susanne; Sztrókay-Gaul, Anikó; Herzen, Julia; Mayr, Doris; Hellerhoff, Karin; Pfeiffer, Franz

    2015-01-01

    Phase-contrast mammography using laboratory X-ray sources is a promising approach to overcome the relatively low sensitivity and specificity of clinical, absorption-based screening. Current research is mostly centered on identifying potential diagnostic benefits arising from phase-contrast and dark-field mammography and benchmarking the latter with conventional state-of-the-art imaging methods. So far, little effort has been made to adjust this novel imaging technique to clinical needs. In this article, we address the key points for a successful implementation to a clinical routine in the near future and present the very first dose-compatible and rapid scan-time phase-contrast mammograms of both a freshly dissected, cancer-bearing mastectomy specimen and a mammographic accreditation phantom. PMID:26110618

  9. Improving Secondary Ion Mass Spectrometry Image Quality with Image Fusion

    PubMed Central

    Tarolli, Jay G.; Jackson, Lauren M.; Winograd, Nicholas

    2014-01-01

    The spatial resolution of chemical images acquired with cluster secondary ion mass spectrometry (SIMS) is limited not only by the size of the probe utilized to create the images, but also by detection sensitivity. As the probe size is reduced to below 1 µm, for example, a low signal in each pixel limits lateral resolution due to counting statistics considerations. Although it can be useful to implement numerical methods to mitigate this problem, here we investigate the use of image fusion to combine information from scanning electron microscope (SEM) data with chemically resolved SIMS images. The advantage of this approach is that the higher intensity and, hence, spatial resolution of the electron images can help to improve the quality of the SIMS images without sacrificing chemical specificity. Using a pan-sharpening algorithm, the method is illustrated using synthetic data, experimental data acquired from a metallic grid sample, and experimental data acquired from a lawn of algae cells. The results show that up to an order of magnitude increase in spatial resolution is possible to achieve. A cross-correlation metric is utilized for evaluating the reliability of the procedure. PMID:24912432

  10. Model-based quantification of image quality

    NASA Technical Reports Server (NTRS)

    Hazra, Rajeeb; Miller, Keith W.; Park, Stephen K.

    1989-01-01

    In 1982, Park and Schowengerdt published an end-to-end analysis of a digital imaging system quantifying three principal degradation components: (1) image blur - blurring caused by the acquisition system, (2) aliasing - caused by insufficient sampling, and (3) reconstruction blur - blurring caused by the imperfect interpolative reconstruction. This analysis, which measures degradation as the square of the radiometric error, includes the sample-scene phase as an explicit random parameter and characterizes the image degradation caused by imperfect acquisition and reconstruction together with the effects of undersampling and random sample-scene phases. In a recent paper Mitchell and Netravelli displayed the visual effects of the above mentioned degradations and presented subjective analysis about their relative importance in determining image quality. The primary aim of the research is to use the analysis of Park and Schowengerdt to correlate their mathematical criteria for measuring image degradations with subjective visual criteria. Insight gained from this research can be exploited in the end-to-end design of optical systems, so that system parameters (transfer functions of the acquisition and display systems) can be designed relative to each other, to obtain the best possible results using quantitative measurements.

  11. Soft copy display requirements for digital mammography.

    PubMed

    Hemminger, Bradley M

    2003-09-01

    One of the advantages of digital mammography is to display mammograms on softcopy (electronic displays). Softcopy display of mammography is challenging because of the spatial and contrast resolution demands present in mammograms. We have designed and developed a softcopy mammography display application, Mammoview, which is capable of allowing radiologists to read mammograms as quickly and as accurately as they can on film alternators. We review the studies using Mammoview to elucidate the requirements of a successful softcopy display station. The design and development of the Mammoview softcopy display station are described in this article, and results of several studies using Mammoview are reported, including subjective feedback from Radiological Society of North America (RSNA) conference demonstrations, and clinical studies measuring performance in terms of speed and accuracy. Additional analysis of user interactions and user feedback is used to study the successes and shortcomings of mammography display stations like Mammoview. Overall, radiologist readings using Mammoview have been shown to be as fast and as accurate as readings using mammography film alternators. However, certain parts of the softcopy interface were more successful than their film counterparts, whereas others were less successful. Data analysis of the recorded human-computer interactions for the softcopy component of the clinical trial indicate statistically significant correlations between the difference in review time of softcopy versus alternator readings and three factors: the number of interactions, the reader, and the size of the image being reviewed. The first factor (number of interactions) suggests that simpler interfaces require less time to use; the second factor, the reader, supports previous findings that radiologists vary in how fast they read screening mammography studies; the third, size of image, suggests that the speed of softcopy review is increased relative to film readings when images are significantly larger than the display size. Feedback from radiologists using the system in clinical trials and at demonstration exhibits at RSNA indicated good acceptance of the interface and easy adaptation. Radiologists indicated that they felt comfortable using the interface, and that they would use such a softcopy interface in clinical practice. Finally, preliminary work suggests that the addition of a simple interaction to incorporate computer-aided detection (CAD) results would improve reading accuracy without significantly increasing reader times. PMID:14669064

  12. Enhancement and quality control of GOES images

    NASA Astrophysics Data System (ADS)

    Jentoft-Nilsen, Marit; Palaniappan, Kannappan; Hasler, A. Frederick; Chesters, Dennis

    1996-10-01

    The new generation of Geostationary Operational Environmental Satellites (GOES) have an imager instrument with five multispectral bands of high spatial resolution,and very high dynamic range radiance measurements with 10-bit precision. A wide variety of environmental processes can be observed at unprecedented time scales using the new imager instrument. Quality assurance and feedback to the GOES project office is performed using rapid animation at high magnification, examining differences between successive frames, and applying radiometric and geometric correction algorithms. Missing or corrupted scanline data occur unpredictably due to noise in the ground based receiving system. Smooth high resolution noise-free animations can be recovered using automatic techniques even from scanline scratches affecting more than 25 percent of the dataset. Radiometric correction using the local solar zenith angle was applied to the visible channel to compensate for time- of-day illumination variations to produce gain-compensated movies that appear well-lit from dawn to dusk and extend the interval of useful image observations by more than two hours. A time series of brightness histograms displays some subtle quality control problems in the GOES channels related to rebinning of the radiance measurements. The human visual system is sensitive to only about half of the measured 10- bit dynamic range in intensity variations, at a given point in a monochrome image. In order to effectively use the additional bits of precision and handle the high data rate, new enhancement techniques and visualization tools were developed. We have implemented interactive image enhancement techniques to selectively emphasize different subranges of the 10-bits of intensity levels. Improving navigational accuracy using registration techniques and geometric correction of scanline interleaving errors is a more difficult problem that is currently being investigated.

  13. Hyperspectral and multispectral imaging for evaluating food safety and quality

    Technology Transfer Automated Retrieval System (TEKTRAN)

    Spectral imaging technologies have been developed rapidly during the past decade. This paper presents hyperspectral and multispectral imaging technologies in the area of food safety and quality evaluation, with an introduction, demonstration, and summarization of the spectral imaging techniques avai...

  14. Objective quality assessment for multiexposure multifocus image fusion.

    PubMed

    Hassen, Rania; Wang, Zhou; Salama, Magdy M A

    2015-09-01

    There has been a growing interest in image fusion technologies, but how to objectively evaluate the quality of fused images has not been fully understood. Here, we propose a method for objective quality assessment of multiexposure multifocus image fusion based on the evaluation of three key factors of fused image quality: 1) contrast preservation; 2) sharpness; and 3) structure preservation. Subjective experiments are conducted to create an image fusion database, based on which, performance evaluation shows that the proposed fusion quality index correlates well with subjective scores, and gives a significant improvement over the existing fusion quality measures. PMID:25935035

  15. Mayo Mammography Health Study

    Cancer.gov

    The Mayo Mammography Health Study (MMHS) is a prospective cohort comprised of 19,924 women ages 35 and over, living in the tri-state region surrounding the Mayo Clinic (Minnesota, Iowa, and Wisconsin), without a history of breast cancer, who were scheduled for a screening mammogram at the Mayo Clinic between October 2003 and September 2006. All women had a 4-view screening mammogram at the time of enrollment and completed a self-administered questionnaire.

  16. On pictures and stuff: image quality and material appearance

    NASA Astrophysics Data System (ADS)

    Ferwerda, James A.

    2014-02-01

    Realistic images are a puzzle because they serve as visual representations of objects while also being objects themselves. When we look at an image we are able to perceive both the properties of the image and the properties of the objects represented by the image. Research on image quality has typically focused improving image properties (resolution, dynamic range, frame rate, etc.) while ignoring the issue of whether images are serving their role as visual representations. In this paper we describe a series of experiments that investigate how well images of different quality convey information about the properties of the objects they represent. In the experiments we focus on the effects that two image properties (contrast and sharpness) have on the ability of images to represent the gloss of depicted objects. We found that different experimental methods produced differing results. Specifically, when the stimulus images were presented using simultaneous pair comparison, observers were influenced by the surface properties of the images and conflated changes in image contrast and sharpness with changes in object gloss. On the other hand, when the stimulus images were presented sequentially, observers were able to disregard the image plane properties and more accurately match the gloss of the objects represented by the different quality images. These findings suggest that in understanding image quality it is useful to distinguish between quality of the imaging medium and the quality of the visual information represented by that medium.

  17. Measuring multivariate subjective image quality for still and video cameras and image processing system components

    NASA Astrophysics Data System (ADS)

    Nyman, Göte; Leisti, Tuomas; Lindroos, Paul; Radun, Jenni; Suomi, Sini; Virtanen, Toni; Olives, Jean-Luc; Oja, Joni; Vuori, Tero

    2008-01-01

    The subjective quality of an image is a non-linear product of several, simultaneously contributing subjective factors such as the experienced naturalness, colorfulness, lightness, and clarity. We have studied subjective image quality by using a hybrid qualitative/quantitative method in order to disclose relevant attributes to experienced image quality. We describe our approach in mapping the image quality attribute space in three cases: still studio image, video clips of a talking head and moving objects, and in the use of image processing pipes for 15 still image contents. Naive observers participated in three image quality research contexts in which they were asked to freely and spontaneously describe the quality of the presented test images. Standard viewing conditions were used. The data shows which attributes are most relevant for each test context, and how they differentiate between the selected image contents and processing systems. The role of non-HVS based image quality analysis is discussed.

  18. Dose Reduction in Automatic Optimization Parameter of Full Field Digital Mammography: Breast Phantom Study

    PubMed Central

    Ko, Myung-Su; Cha, Joo Hee; Shin, Hee Jung; Kim, Jeoung Hyun; Kim, Min Jeong

    2013-01-01

    Purpose We evaluated the impact of three automatic optimization of parameters (AOP) modes of digital mammography on the dose and image quality. Methods Computerized Imaging Reference Systems phantoms were used. A total of 12 phantoms with different thickness and glandularity were imaged. We analyzed the average glandular dose (AGD) and entrance surface exposure (ESE) of 12 phantoms imaged by digital mammography in three modes of AOP; namely standard mode (STD), contrast mode (CNT), and dose mode (DOSE). Moreover, exposure factors including kVp, mAs, and target/filter combination were evaluated. To evaluate the quality of the obtained digital image, two radiologists independently counted the objects of the phantoms. Results According to the AOP modes, the score of masses and specks was sorted as CNT>STD=DOSE. There was no difference in the score of fiber among the three modes. The score of image preference was sorted as CNT>STD>DOSE. The AGD, ESE, and mAs were sorted as CNT>STD>DOSE. The kVp was sorted as CNT=STD>DOSE. The score of all test objects in the phantom image was on a downtrend with increasing breast thickness. The score of masses was different among the three groups; 20-21%>30%>50% glandularity. The score of specks was sorted as 20-21%=30%>50% glandularity. The score of fibers was sorted as 30%>20-21%=50% glandularity. The score of image preference was not different among the three glandularity groups. The AGD, ESE, kVp, and mAs were correlated with breast thickness, but not correlated with glandularity. Conclusion The DOSE mode offers significant improvement (19.1-50%) in dose over the other two modes over a range of breast thickness and breast glandularity with acceptable image quality. Owning knowledge of the three AOP modes may reduce unnecessary radiation exposure by utilizing the proper mode according to its purpose. PMID:23593088

  19. Bi-Directional X-Ray Phase-Contrast Mammography

    PubMed Central

    Scherer, Kai; Birnbacher, Lorenz; Chabior, Michael; Herzen, Julia; Mayr, Doris; Grandl, Susanne; Sztrókay-Gaul, Anikó; Hellerhoff, Karin; Bamberg, Fabian; Pfeiffer, Franz

    2014-01-01

    Phase-contrast x-ray imaging is a promising improvement of conventional absorption-based mammography for early tumor detection. This potential has been demonstrated recently, utilizing structured gratings to obtain differential phase and dark-field scattering images. However, the inherently anisotropic imaging sensitivity of the proposed mono-directional approach yields only insufficient diagnostic information, and has low diagnostic sensitivity to highly oriented structures. To overcome these limitations, we present a two-directional x-ray phase-contrast mammography approach and demonstrate its advantages by applying it to a freshly dissected, cancerous mastectomy breast specimen. We illustrate that the two-directional scanning procedure overcomes the insufficient diagnostic value of a single scan, and reliably detects tumor structures, independently from their orientation within the breast. Our results indicate the indispensable diagnostic necessity and benefit of a multi-directional approach for x-ray phase-contrast mammography. PMID:24824594

  20. Image quality metrics for optical coherence angiography

    PubMed Central

    Lozzi, Andrea; Agrawal, Anant; Boretsky, Adam; Welle, Cristin G.; Hammer, Daniel X.

    2015-01-01

    We characterized image quality in optical coherence angiography (OCA) en face planes of mouse cortical capillary network in terms of signal-to-noise ratio (SNR) and Weber contrast (Wc) through a novel mask-based segmentation method. The method was used to compare two adjacent B-scan processing algorithms, (1) average absolute difference (AAD) and (2) standard deviation (SD), while varying the number of lateral cross-sections acquired (also known as the gate length, N). AAD and SD are identical at N = 2 and exhibited similar image quality for N<10. However, AAD is relatively less susceptible to bulk tissue motion artifact than SD. SNR and Wc were 15% and 35% higher for AAD from N = 25 to 100. In addition data sets were acquired with two objective lenses with different magnifications to quantify the effect of lateral resolution on fine capillary detection. The lower power objective yielded a significant mean broadening of 17% in Full Width Half Maximum (FWHM) diameter. These results may guide study and device designs for OCA capillary and blood flow quantification. PMID:26203372

  1. Image simulation and a model of noise power spectra across a range of mammographic beam qualities

    SciTech Connect

    Mackenzie, Alistair Dance, David R.; Young, Kenneth C.; Diaz, Oliver

    2014-12-15

    Purpose: The aim of this work is to create a model to predict the noise power spectra (NPS) for a range of mammographic radiographic factors. The noise model was necessary to degrade images acquired on one system to match the image quality of different systems for a range of beam qualities. Methods: Five detectors and x-ray systems [Hologic Selenia (ASEh), Carestream computed radiography CR900 (CRc), GE Essential (CSI), Carestream NIP (NIPc), and Siemens Inspiration (ASEs)] were characterized for this study. The signal transfer property was measured as the pixel value against absorbed energy per unit area (E) at a reference beam quality of 28 kV, Mo/Mo or 29 kV, W/Rh with 45 mm polymethyl methacrylate (PMMA) at the tube head. The contributions of the three noise sources (electronic, quantum, and structure) to the NPS were calculated by fitting a quadratic at each spatial frequency of the NPS against E. A quantum noise correction factor which was dependent on beam quality was quantified using a set of images acquired over a range of radiographic factors with different thicknesses of PMMA. The noise model was tested for images acquired at 26 kV, Mo/Mo with 20 mm PMMA and 34 kV, Mo/Rh with 70 mm PMMA for three detectors (ASEh, CRc, and CSI) over a range of exposures. The NPS were modeled with and without the noise correction factor and compared with the measured NPS. A previous method for adapting an image to appear as if acquired on a different system was modified to allow the reference beam quality to be different from the beam quality of the image. The method was validated by adapting the ASEh flat field images with two thicknesses of PMMA (20 and 70 mm) to appear with the imaging characteristics of the CSI and CRc systems. Results: The quantum noise correction factor rises with higher beam qualities, except for CR systems at high spatial frequencies, where a flat response was found against mean photon energy. This is due to the dominance of secondary quantum noise in CR. The use of the quantum noise correction factor reduced the difference from the model to the real NPS to generally within 4%. The use of the quantum noise correction improved the conversion of ASEh image to CRc image but had no difference for the conversion to CSI images. Conclusions: A practical method for estimating the NPS at any dose and over a range of beam qualities for mammography has been demonstrated. The noise model was incorporated into a methodology for converting an image to appear as if acquired on a different detector. The method can now be extended to work for a wide range of beam qualities and can be applied to the conversion of mammograms.

  2. Mammography dosimetry using an in-house developed polymethyl methacrylate phantom.

    PubMed

    Sharma, Reena; Sharma, Sunil Dutt; Mayya, Y S; Chourasiya, G

    2012-08-01

    Phantom-based measurements in mammography are well-established for quality assurance (QA) and quality control (QC) procedures involving equipment performance and comparisons of X-ray machines. Polymethyl methacrylate (PMMA) is among the best suitable materials for simulation of the breast. For carrying out QA/QC exercises in India, a mammographic PMMA phantom with engraved slots for keeping thermoluminescence dosemeters (TLD) has been developed. The radiation transmission property of the developed phantom was compared with the commercially available phantoms for verifying its suitability for mammography dosimetry. The breast entrance exposure (BEE), mean glandular dose (MGD), percentage depth dose (PDD), percentage surface dose distribution (PSDD), calibration testing of automatic exposure control (AEC) and density control function of a mammography machine were measured using this phantom. MGD was derived from the measured BEE following two different methodologies and the results were compared. The PDD and PSDD measurements were carried out using LiF: Mg, Cu, P chips. The in-house phantom was found comparable with the commercially available phantoms. The difference in the MGD values derived using two different methods were found in the range of 17.5-32.6 %. Measured depth ranges in the phantom lie between 0.32 and 0.40 cm for 75 % depth dose, 0.73 and 0.92 cm for 50 % depth dose, and 1.54 and 1.78 cm for 25 % depth dose. Higher PSDD value was observed towards chest wall edge side of the phantom, which is due to the orientation of cathode-anode axis along the chest wall to the nipple direction. Results obtained for AEC configuration testing shows that the observed mean optical density (O.D) of the phantom image was 1.59 and O.D difference for every successive increase in thickness of the phantom was within±0.15 O.D. Under density control function testing, at -2 and -1 density settings, the variation in film image O.D was within±0.15 O.D of the normal density setting '0' and at +2 and +1 density setting, it was observed to be within±0.30 O.D. This study indicates that the locally made PMMA TLD slot phantom can be used to measure various mammography QC parameters which are essentially required for better outcomes in mammography. PMID:22232773

  3. [The diagnostic accuracy of a digital mammography system with photostimulable storage phosphors used with automatic reading].

    PubMed

    Panizza, P; Rodighiero, M G; De Gaspari, A; Tacchini, S; Camalori, M; Del Maschio, A

    1996-01-01

    Even though digital mammography might potentially yield major advantages in management, biology and diagnosis, state-of-the-art digitalization is still in a clinical experimental phase. To assess the diagnostic accuracy of a digital mammography system with storage phosphors, we analyzed 320 digital and 320 conventional mammograms acquired in lateral-oblique projection in the same patients. Digital mammography capabilities in identifying and characterizing breast lesions were compared with those of conventional mammography; the presence/absence of lesions and their benign/malignant nature were investigated. Complete conventional mammography was our gold standard. The digital system did not miss any malignant lesion but it did miss some benign lesions (focal masses) which had been depicted with conventional mammography, especially small low-contrast opacities. Microcalcifications were better depicted on digital images which showed, at the same time, tissues of different density, thanks to their wider dynamic range. Digital mammography yielded a false positive result on a cluster of microcalcifications, because its spatial resolution is lower. The diagnostic yield of digital mammography was poorer only in the detection of small low-contrast lesions and in the characterization of microcalcifications. To conclude, in our experience, the storage phosphor system seems to be suitable for clinical mammography, but only with careful monitoring; in contrast, we think it is not yet suitable for screening purposes. PMID:8614730

  4. Image Quality Characteristics of Handheld Display Devices for Medical Imaging

    PubMed Central

    Yamazaki, Asumi; Liu, Peter; Cheng, Wei-Chung; Badano, Aldo

    2013-01-01

    Handheld devices such as mobile phones and tablet computers have become widespread with thousands of available software applications. Recently, handhelds are being proposed as part of medical imaging solutions, especially in emergency medicine, where immediate consultation is required. However, handheld devices differ significantly from medical workstation displays in terms of display characteristics. Moreover, the characteristics vary significantly among device types. We investigate the image quality characteristics of various handheld devices with respect to luminance response, spatial resolution, spatial noise, and reflectance. We show that the luminance characteristics of the handheld displays are different from those of workstation displays complying with grayscale standard target response suggesting that luminance calibration might be needed. Our results also demonstrate that the spatial characteristics of handhelds can surpass those of medical workstation displays particularly for recent generation devices. While a 5 mega-pixel monochrome workstation display has horizontal and vertical modulation transfer factors of 0.52 and 0.47 at the Nyquist frequency, the handheld displays released after 2011 can have values higher than 0.63 at the respective Nyquist frequencies. The noise power spectra for workstation displays are higher than 1.2×10?5 mm2 at 1 mm?1, while handheld displays have values lower than 3.7×10?6 mm2. Reflectance measurements on some of the handheld displays are consistent with measurements for workstation displays with, in some cases, low specular and diffuse reflectance coefficients. The variability of the characterization results among devices due to the different technological features indicates that image quality varies greatly among handheld display devices. PMID:24236113

  5. X-ray light valve (XLV): a novel detectors' technology for digital mammography

    NASA Astrophysics Data System (ADS)

    Marcovici, Sorin; Sukhovatkin, Vlad; Oakham, Peter

    2014-03-01

    A novel method, based on X-ray Light Valve (XLV) technology, is proposed for making good image quality yet inexpensive flat panel detectors for digital mammography. The digital mammography markets, particularly in the developing countries, demand quality machines at substantially lower prices than the ones available today. Continuous pressure is applied on x-ray detectors' manufacturers to reduce the flat panel detectors' prices. XLV presents a unique opportunity to achieve the needed price - performance characteristics for direct conversion, x-ray detectors. The XLV based detectors combine the proven, superior, spatial resolution of a-Se with the simplicity and low cost of liquid crystals and optical scanning. The x-ray quanta absorbed by a 200 ?m a-Se produce electron - hole pairs that move under an electric field to the top and bottom of a-Se layer. This 2D charge distribution creates at the interface with the liquid crystals a continuous (analog) charge image corresponding to the impinging radiation's information. Under the influence of local electrical charges next to them, the liquid crystals twist proportionally to the charges and vary their light reflectivity. A scanning light source illuminates the liquid crystals while an associated, pixilated photo-detector, having a 42 ?m pixel size, captures the light reflected by the liquid crystals and converts it in16 bit words that are transmitted to the machine for image processing and display. The paper will describe a novel XLV, 25 cm x 30 cm, flat panel detector structure and its underlying physics as well as its preliminary performance measured on several engineering prototypes. In particular, the paper will present the results of measuring XLV detectors' DQE, MTF, dynamic range, low contrast resolution and dynamic behavior. Finally, the paper will introduce the new, low cost, XLV detector based, digital mammography machine under development at XLV Diagnostics Inc.

  6. The effect of breast compression on mass conspicuity in digital mammography

    SciTech Connect

    Saunders, Robert S. Jr; Samei, Ehsan

    2008-10-15

    This study analyzed how the inherent quality of diagnostic information in digital mammography could be affected by breast compression. A digital mammography system was modeled using a Monte Carlo algorithm based on the Penelope program, which has been successfully used to model several medical imaging systems. First, the Monte Carlo program was validated against previous measurements and simulations. Once validated, the Monte Carlo software modeled a digital mammography system by tracking photons through a voxelized software breast phantom, containing anatomical structures and breast masses, and following photons until they were absorbed by a selenium-based flat-panel detector. Simulations were performed for two compression conditions (standard compression and 12.5% reduced compression) and three photon flux conditions (constant flux, constant detector signal, and constant glandular dose). The results showed that reduced compression led to higher scatter fractions, as expected. For the constant photon flux condition, decreased compression also reduced glandular dose. For constant glandular dose, the SdNR for a 4 cm breast was 0.60{+-}0.11 and 0.62{+-}0.11 under standard and reduced compressions, respectively. For the 6 cm case with constant glandular dose, the SdNR was 0.50{+-}0.11 and 0.49{+-}0.10 under standard and reduced compressions, respectively. The results suggest that if a particular imaging system can handle an approximately 10% increase in total tube output and 10% decrease in detector signal, breast compression can be reduced by about 12% in terms of breast thickness with little impact on image quality or dose.

  7. Quality assessment for spectral domain optical coherence tomography (OCT) images.

    PubMed

    Liu, Shuang; Paranjape, Amit S; Elmaanaoui, Badr; Dewelle, Jordan; Rylander, H Grady; Markey, Mia K; Milner, Thomas E

    2009-01-01

    Retinal nerve fiber layer (RNFL) thickness, a measure of glaucoma progression, can be measured in images acquired by spectral domain optical coherence tomography (OCT). The accuracy of RNFL thickness estimation, however, is affected by the quality of the OCT images. In this paper, a new parameter, signal deviation (SD), which is based on the standard deviation of the intensities in OCT images, is introduced for objective assessment of OCT image quality. Two other objective assessment parameters, signal to noise ratio (SNR) and signal strength (SS), are also calculated for each OCT image. The results of the objective assessment are compared with subjective assessment. In the subjective assessment, one OCT expert graded the image quality according to a three-level scale (good, fair, and poor). The OCT B-scan images of the retina from six subjects are evaluated by both objective and subjective assessment. From the comparison, we demonstrate that the objective assessment successfully differentiates between the acceptable quality images (good and fair images) and poor quality OCT images as graded by OCT experts. We evaluate the performance of the objective assessment under different quality assessment parameters and demonstrate that SD is the best at distinguishing between fair and good quality images. The accuracy of RNFL thickness estimation is improved significantly after poor quality OCT images are rejected by automated objective assessment using the SD, SNR, and SS. PMID:20431701

  8. Quality assessment for spectral domain optical coherence tomography (OCT) images

    NASA Astrophysics Data System (ADS)

    Liu, Shuang; Paranjape, Amit S.; Elmaanaoui, Badr; Dewelle, Jordan; Rylander, H. Grady, III; Markey, Mia K.; Milner, Thomas E.

    2009-02-01

    Retinal nerve fiber layer (RNFL) thickness, a measure of glaucoma progression, can be measured in images acquired by spectral domain optical coherence tomography (OCT). The accuracy of RNFL thickness estimation, however, is affected by the quality of the OCT images. In this paper, a new parameter, signal deviation (SD), which is based on the standard deviation of the intensities in OCT images, is introduced for objective assessment of OCT image quality. Two other objective assessment parameters, signal to noise ratio (SNR) and signal strength (SS), are also calculated for each OCT image. The results of the objective assessment are compared with subjective assessment. In the subjective assessment, one OCT expert graded the image quality according to a three-level scale (good, fair, and poor). The OCT B-scan images of the retina from six subjects are evaluated by both objective and subjective assessment. From the comparison, we demonstrate that the objective assessment successfully differentiates between the acceptable quality images (good and fair images) and poor quality OCT images as graded by OCT experts. We evaluate the performance of the objective assessment under different quality assessment parameters and demonstrate that SD is the best at distinguishing between fair and good quality images. The accuracy of RNFL thickness estimation is improved significantly after poor quality OCT images are rejected by automated objective assessment using the SD, SNR, and SS.

  9. Motion correction via nonrigid coregistration of dynamic MR mammography series

    NASA Astrophysics Data System (ADS)

    Krol, Andrzej; Magri, Alphonso; Unlu, Mehmet; Feiglin, David; Lipson, Edward; Mandel, James; Tillapaugh-Fay, Gwen; Lee, Wei; Coman, Ioana; Szeverenyi, Nikolaus M.

    2006-03-01

    The objectives of this investigation are to improve quality of subtraction MR breast images and improve accuracy of time-signal intensity curves (TSIC) related to local contrast-agent concentration in dynamic MR mammography. The patients, with up to nine fiducial skin markers (FSMs) taped to each breast, were prone with both breasts suspended into a single well that housed the receiver coil. After a preliminary scan, paramagnetic contrast agent gadopentate digmeglumine (Gd) was delivered intravenously, followed by physiological saline. The field of view was centered over the breasts. We used a gradient recalled echo (GRE) technique for pre-Gd baseline, and five more measurements at 60s intervals. Centroids were determined for corresponding FSMs visible on pre-Gd and any post-Gd images. This was followed by segmentation of breast surfaces in all dynamic-series images, and meshing of all post-Gd breast images. Tetrahedral volume and triangular surface elements were used to construct a finite element method (FEM) model. We used ANSYS TM software and an analogy between orthogonal components of the displacement field and the temperature differences in steady-state heat transfer (SSHT) in solids. The floating images were warped to a fixed image using an appropriate shape function for interpolation from mesh nodes to voxels. To reduce any residual misregistration, we performed surface matching between the previously warped floating image and the target image. Our method of motion correction via nonrigid coregistration yielded excellent differential-image series that clearly revealed lesions not visible in unregistered differential-image series. Further, it produced clinically useful maximum intensity projection (MIP) 3D images.

  10. The influence of statistical variations on image quality

    NASA Astrophysics Data System (ADS)

    Hultgren, Bror; Hertel, Dirk; Bullitt, Julian

    2006-01-01

    For more than thirty years imaging scientists have constructed metrics to predict psychovisually perceived image quality. Such metrics are based on a set of objectively measurable basis functions such as Noise Power Spectrum (NPS), Modulation Transfer Function (MTF), and characteristic curves of tone and color reproduction. Although these basis functions constitute a set of primitives that fully describe an imaging system from the standpoint of information theory, we found that in practical imaging systems the basis functions themselves are determined by system-specific primitives, i.e. technology parameters. In the example of a printer, MTF and NPS are largely determined by dot structure. In addition MTF is determined by color registration, and NPS by streaking and banding. Since any given imaging system is only a single representation of a class of more or less identical systems, the family of imaging systems and the single system are not described by a unique set of image primitives. For an image produced by a given imaging system, the set of image primitives describing that particular image will be a singular instantiation of the underlying statistical distribution of that primitive. If we know precisely the set of imaging primitives that describe the given image we should be able to predict its image quality. Since only the distributions are known, we can only predict the distribution in image quality for a given image as produced by the larger class of 'identical systems'. We will demonstrate the combinatorial effect of the underlying statistical variations in the image primitives on the objectively measured image quality of a population of printers as well as on the perceived image quality of a set of test images. We also will discuss the choice of test image sets and impact of scene content on the distribution of perceived image quality.

  11. Beyond MQSA: Measuring the quality of breast cancer screening programs

    PubMed Central

    Rauscher, Garth H.; Murphy, Anne Marie; Orsi, Jennifer M.; Dupuy, Danielle M.; Grabler, Paula M.; Weldon, Christine B.

    2014-01-01

    OBJECTIVE A high quality screening mammography program should find breast cancer when it exists, when it’s small, and ensure that suspicious findings receive prompt follow-up. The Mammography Quality Standards Act (MQSA) guidelines related to tracking outcomes are insufficient for assessing quality of care. We used data from a quality improvement project to determine whether mammography screening facilities could show that they met certain quality benchmarks beyond those required by MQSA. METHODS Participating facilities (N=52) provided aggregate data on screening mammograms conducted in calendar year 2009 and corresponding diagnostic follow-up, including lost to follow-up and timing of diagnostic imaging and biopsy, cancer detection rates, and the proportion of cancers detected as minimal and early stage tumors. RESULTS The percentage of institutions meeting each benchmark varied from 27% to 83%. Facilities with American College of Surgeons or National Comprehensive Cancer Network designation were more likely to meet benchmarks pertaining to cancer detection and early detection, and Disproportionate Share facilities were less likely to meet benchmarks pertaining to timeliness of care. CONCLUSIONS Results suggest a combination of care quality issues and incomplete tracking of patients. To accurately measure quality of the breast cancer screening process, it is critical that there be complete tracking of patients with abnormal screening mammograms so that results can be interpreted solely in terms of quality of care. The Mammography Quality Standards Act guidelines for tracking outcomes and measuring quality indicators should be strengthened to better assess quality of care. PMID:24261339

  12. Using short-wave infrared imaging for fruit quality evaluation

    NASA Astrophysics Data System (ADS)

    Zhang, Dong; Lee, Dah-Jye; Desai, Alok

    2013-12-01

    Quality evaluation of agricultural and food products is important for processing, inventory control, and marketing. Fruit size and surface quality are two important quality factors for high-quality fruit such as Medjool dates. Fruit size is usually measured by length that can be done easily by simple image processing techniques. Surface quality evaluation on the other hand requires more complicated design, both in image acquisition and image processing. Skin delamination is considered a major factor that affects fruit quality and its value. This paper presents an efficient histogram analysis and image processing technique that is designed specifically for real-time surface quality evaluation of Medjool dates. This approach, based on short-wave infrared imaging, provides excellent image contrast between the fruit surface and delaminated skin, which allows significant simplification of image processing algorithm and reduction of computational power requirements. The proposed quality grading method requires very simple training procedure to obtain a gray scale image histogram for each quality level. Using histogram comparison, each date is assigned to one of the four quality levels and an optimal threshold is calculated for segmenting skin delamination areas from the fruit surface. The percentage of the fruit surface that has skin delamination can then be calculated for quality evaluation. This method has been implemented and used for commercial production and proven to be efficient and accurate.

  13. Quality Prediction of Asymmetrically Distorted Stereoscopic 3D Images.

    PubMed

    Wang, Jiheng; Rehman, Abdul; Zeng, Kai; Wang, Shiqi; Wang, Zhou

    2015-11-01

    Objective quality assessment of distorted stereoscopic images is a challenging problem, especially when the distortions in the left and right views are asymmetric. Existing studies suggest that simply averaging the quality of the left and right views well predicts the quality of symmetrically distorted stereoscopic images, but generates substantial prediction bias when applied to asymmetrically distorted stereoscopic images. In this paper, we first build a database that contains both single-view and symmetrically and asymmetrically distorted stereoscopic images. We then carry out a subjective test, where we find that the quality prediction bias of the asymmetrically distorted images could lean toward opposite directions (overestimate or underestimate), depending on the distortion types and levels. Our subjective test also suggests that eye dominance effect does not have strong impact on the visual quality decisions of stereoscopic images. Furthermore, we develop an information content and divisive normalization-based pooling scheme that improves upon structural similarity in estimating the quality of single-view images. Finally, we propose a binocular rivalry-inspired multi-scale model to predict the quality of stereoscopic images from that of the single-view images. Our results show that the proposed model, without explicitly identifying image distortion types, successfully eliminates the prediction bias, leading to significantly improved quality prediction of the stereoscopic images. PMID:26087491

  14. LANDSAT-4 image data quality analysis

    NASA Technical Reports Server (NTRS)

    Anuta, P. E. (principal investigator)

    1982-01-01

    Work done on evaluating the geometric and radiometric quality of early LANDSAT-4 sensor data is described. Band to band and channel to channel registration evaluations were carried out using a line correlator. Visual blink comparisons were run on an image display to observe band to band registration over 512 x 512 pixel blocks. The results indicate a .5 pixel line misregistration between the 1.55 to 1.75, 2.08 to 2.35 micrometer bands and the first four bands. Also a four 30M line and column misregistration of the thermal IR band was observed. Radiometric evaluation included mean and variance analysis of individual detectors and principal components analysis. Results indicate that detector bias for all bands is very close or within tolerance. Bright spots were observed in the thermal IR band on an 18 line by 128 pixel grid. No explanation for this was pursued. The general overall quality of the TM was judged to be very high.

  15. Food quality assessment by NIR hyperspectral imaging

    NASA Astrophysics Data System (ADS)

    Whitworth, Martin B.; Millar, Samuel J.; Chau, Astor

    2010-04-01

    Near infrared reflectance (NIR) spectroscopy is well established in the food industry for rapid compositional analysis of bulk samples. NIR hyperspectral imaging provides new opportunities to measure the spatial distribution of components such as moisture and fat, and to identify and measure specific regions of composite samples. An NIR hyperspectral imaging system has been constructed for food research applications, incorporating a SWIR camera with a cooled 14 bit HgCdTe detector and N25E spectrograph (Specim Ltd, Finland). Samples are scanned in a pushbroom mode using a motorised stage. The system has a spectral resolution of 256 pixels covering a range of 970-2500 nm and a spatial resolution of 320 pixels covering a swathe adjustable from 8 to 300 mm. Images are acquired at a rate of up to 100 lines s-1, enabling samples to be scanned within a few seconds. Data are captured using SpectralCube software (Specim) and analysed using ENVI and IDL (ITT Visual Information Solutions). Several food applications are presented. The strength of individual absorbance bands enables the distribution of particular components to be assessed. Examples are shown for detection of added gluten in wheat flour and to study the effect of processing conditions on fat distribution in chips/French fries. More detailed quantitative calibrations have been developed to study evolution of the moisture distribution in baguettes during storage at different humidities, to assess freshness of fish using measurements of whole cod and fillets, and for prediction of beef quality by identification and separate measurement of lean and fat regions.

  16. Retinal Image Quality during Accommodation in Adult Myopic Eyes

    PubMed Central

    Sreenivasan, Vidhyapriya; Aslakson, Emily; Kornaus, Andrew; Thibos, Larry N.

    2014-01-01

    Purpose Reduced retinal image contrast produced by accommodative lag is implicated with myopia development. Here, we measure accommodative error and retinal image quality from wavefront aberrations in myopes and emmetropes when they perform visually demanding and naturalistic tasks. Methods Wavefront aberrations were measured in 10 emmetropic and 11 myopic adults at three distances (100, 40, and 20 cm) while performing four tasks (monocular acuity, binocular acuity, reading, and movie watching). For the acuity tasks, measurements of wavefront error were obtained near the end point of the acuity experiment. Refractive state was defined as the target vergence that optimizes image quality using a visual contrast metric (VSMTF) computed from wavefront errors. Results Accommodation was most accurate (and image quality best) during binocular acuity whereas accommodation was least accurate (and image quality worst) while watching a movie. When viewing distance was reduced, accommodative lag increased and image quality (as quantified by VSMTF) declined for all tasks in both refractive groups. For any given viewing distance, computed image quality was consistently worse in myopes than in emmetropes, more so for the acuity than for reading/movie watching. Although myopes showed greater lags and worse image quality for the acuity experiments compared to emmetropes, acuity was not measurably worse in myopes compared to emmetropes. Conclusions Retinal image quality present when performing a visually demanding task (e.g., during clinical examination) is likely to be greater than for less demanding tasks (e.g., reading/movie watching). Although reductions in image quality lead to reductions in acuity, the image quality metric VSMTF is not necessarily an absolute indicator of visual performance because myopes achieved slightly better acuity than emmetropes despite showing greater lags and worse image quality. Reduced visual contrast in myopes compared to emmetropes is consistent with theories of myopia progression that point to image contrast as an inhibitory signal for ocular growth. PMID:24152885

  17. Perceptual Quality Assessment for Multi-Exposure Image Fusion.

    PubMed

    Ma, Kede; Zeng, Kai; Wang, Zhou

    2015-11-01

    Multi-exposure image fusion (MEF) is considered an effective quality enhancement technique widely adopted in consumer electronics, but little work has been dedicated to the perceptual quality assessment of multi-exposure fused images. In this paper, we first build an MEF database and carry out a subjective user study to evaluate the quality of images generated by different MEF algorithms. There are several useful findings. First, considerable agreement has been observed among human subjects on the quality of MEF images. Second, no single state-of-the-art MEF algorithm produces the best quality for all test images. Third, the existing objective quality models for general image fusion are very limited in predicting perceived quality of MEF images. Motivated by the lack of appropriate objective models, we propose a novel objective image quality assessment (IQA) algorithm for MEF images based on the principle of the structural similarity approach and a novel measure of patch structural consistency. Our experimental results on the subjective database show that the proposed model well correlates with subjective judgments and significantly outperforms the existing IQA models for general image fusion. Finally, we demonstrate the potential application of the proposed model by automatically tuning the parameters of MEF algorithms. PMID:26068317

  18. Blind quality assessment of multi-focus image fusion algorithms

    NASA Astrophysics Data System (ADS)

    Nava, Rodrigo; Escalante-Ramírez, Boris; Cristóbal, Gabriel

    2010-05-01

    At present time, image fusion is widely recognized as an important aspect of information processing. It consists of combining information originated from several sources in order to improve the decision making process. In particular, multi-focus image fusion combines images that depict the same scene but they are not in-focus everywhere. The task seeks to reconstruct an image as sharp as possible by preserving in-focus areas while discarding blurred areas. The quality of fused images is of fundamental importance. Many objective quality metrics for image fusion have been proposed. However, the evaluation of fused images is still a difficult task, especially because there is no reference image to compare with. Blind image quality assessment refers to the problem of evaluating the visual quality of an image without any reference. In this paper, we describe a blind image fusion quality assessment procedure based on the use of mutual information (MI). This procedure is concise and explicit and will be useful in scenarios where the absence of a reference image can hamper the assessment of the results. Furthermore, several image fusion algorithms have been rated and they have shown that our metric is compliant with subjective evaluations. Consequently, it can be used to compare different image fusion methods or to optimize the parameter settings for a given fusion algorithm.

  19. Hidden costs of poor image quality: a radiologist's perspective.

    PubMed

    Reiner, Bruce I

    2014-10-01

    Although image quality is a well-recognized component in the successful delivery of medical imaging services, it has arguably declined over the past decade owing to several technical, economic, cultural, and geographic factors. To improve quality, the radiologist community must take a more proactive role in image quality analysis and optimization; these require analysis of not just the single step of image acquisition but the entire imaging chain. Radiologists can benefit through improved report accuracy, diagnostic confidence, and workflow efficiency. The derived data-driven analyses offer an objective means for provider performance analysis, which can help combat commoditization trends and self-referral by nonradiologist providers. PMID:24889471

  20. Visibility of tumor-like details in inline phase contrast mammography using quasimonochromatic X-ray sources

    NASA Astrophysics Data System (ADS)

    Golosio, B.; Delogu, P.; Zanette, I.; Oliva, P.; Stefanini, A.; Stegel, G.; Carpinelli, M.

    2009-09-01

    A new generation of quasimonochromatic high-flux X-ray sources, based on the X-ray radiation produced through Compton scattering between an electron beam and a laser beam, is under development. One of the possible applications of this source is inline phase contrast mammography, based on the observation of the edge-enhancement effect that can be observed at the border of structures inside the breast in images produced using a partially or totally coherent X-ray beam. In this work we present the results of a set of simulations of inline phase contrast mammography using typical inverse Compton scattering sources parameters. The simulated sample was a tumour-like mass having spherical shape, diameter between 200 ?m and 5 mm, placed inside a breast-like matrix, 4 cm thick, and a standard composition of 50% glandular tissue and 50% adipose tissue. We discuss the minimum requirements for mammography using inverse Compton scattering sources and we discuss how the working parameters of the experimental setup (focal spot size, source-object distance, object-detector distance, detector point spread function, mean energy, energy bandwidth) affect the image quality, and specifically the edge-enhancement visibility. In particular, we show that the energy bandwidth does not significantly affect the visibility up to several keV. On the other hand, the edge-enhancement visibility depends substantially on other parameters such as source-object distance and detector point spread function.

  1. Reducing disparities in mammography-use in a multicultural population in Israel

    PubMed Central

    Baron-Epel, Orna; Friedman, Nurit; Lernau, Omri

    2009-01-01

    Background In the past mammography-use has been reported to be low in Israel compared to other western countries. The objectives of this study were (1) to assess the increase in mammography-use during the years 2002 to 2007 in four population groups in Maccabi Healthcare Services (MHS), Israel: non-immigrant non-ultraorthodox, ultraorthodox, and immigrant Jewish women and Arab women; (2) to assess ethnic and socioeconomic disparities in mammography-use. Methods A random telephone survey of 1,550 women receiving healthcare services from MHS was performed during May-June 2007. Information from MHS claims-records database regarding mammography-use was obtained for each woman for the period 2002 to 2007. Since mammography-use serves as a quality assurance measure for primary care, MHS sent mail and telephone invitations for mammography to all women since the end of 2004. Results At the beginning of the follow-up period (2002) mammography-use among Jewish non-immigrant non-ultraorthodox and ultraorthodox women was higher than among Arab and Jewish immigrant women. During the 5 year follow-up these disparities decreased significantly. In 2007, mammography-use by Arab women was only slightly lower compared to all groups of Jewish women. In 2007, after adjustment for socioeconomic factors there was only a borderline significant difference between Jewish and Arab women. The socioeconomic variables were not associated with mammography-use in 2002 and 2007 in any of the groups except for marital status in immigrant women in 2002. Conclusion The interventions implemented by MHS may have increased mammography-use in all population groups, decreasing disparities between the groups, however the differences between Jewish and Arab women have not been completely eliminated and indicate a need for further targeted interventions. No significant socioeconomic disparities in mammography-use were observed. PMID:19454004

  2. Searching for the limit of image quality in film radiography

    SciTech Connect

    Vaessen, B.; Perdieus, P.; Florens, R.

    1993-12-31

    Radiographic film image quality in general was, and in most cases still is, considered as a very subjective and rather vague parameter. Yet it is of vital importance to the NDT and related quality control and quality assurance industry. Therefore, lately Agfa has put a major effort into quantifying image quality in an objective, measurable way. It was in the framework of this optimization project, that the authors, based on these new insights in imaging of industrial film systems, strived to search for the limit of the highest achievable image quality. In this paper they report these results. They not only report these results in an academic way, meaning how this highest image quality can be achieved under lab conditions, but also how these same results can be obtained under practical e.g. field-conditions.

  3. The influence of environment temperature on SEM image quality

    NASA Astrophysics Data System (ADS)

    Chen, Li; Liu, Junshan

    2015-07-01

    As the structure dimension goes down to the nano-scale, it often requires a scanning electron microscope (SEM) to provide image magnification up to 100?000??×. However, SEM images at such a high magnification usually suffer from high resolution value and low signal-to-noise ratio, which results in low quality of the SEM image. In this paper, the quality of the SEM image is improved by optimizing the environment temperature. The experimental results indicate that at 100?000??×, the quality of the SEM image is influenced by the environment temperature, whereas at 50?000??× it is not. At 100?000??× the best SEM image quality can be achieved from the environment temperature ranging 292 from 294?K, and the SEM image quality evaluated by the double stimulus continuous quality scale method can increase from grade 1 to grade 5. It is expected that this image quality improving method can be used in routine measurements with ordinary SEMs to get high quality images by optimizing the environment temperature.

  4. The evolution of breast imaging: past to present.

    PubMed

    Joe, Bonnie N; Sickles, Edward A

    2014-11-01

    The practice of breast imaging has transitioned through a wide variety of technologic advances from the early days of direct-exposure film mammography to xeromammography to screen-film mammography to the current era of full-field digital mammography and digital breast tomosynthesis. Along with these technologic advances, organized screening, federal regulations based on the Mammography Quality Standards Act, and the development of the American College of Radiology Breast Imaging Reporting and Data System have helped to shape the specialty of breast imaging. With the development of breast ultrasonography and breast magnetic resonance imaging, both complementary to mammography, additional algorithms for diagnostic workup and screening high-risk subgroups of women have emerged. A substantial part of breast imaging practice these days also involves breast interventional procedures-both percutaneous biopsy to obtain tissue diagnosis and localization procedures to guide surgical excision. This article reviews the evolution of breast imaging starting from a historical perspective and progressing to the present day. PMID:25340437

  5. Limitations to adaptive optics image quality in rodent eyes

    PubMed Central

    Zhou, Xiaolin; Bedggood, Phillip; Metha, Andrew

    2012-01-01

    Adaptive optics (AO) retinal image quality of rodent eyes is inferior to that of human eyes, despite the promise of greater numerical aperture. This paradox challenges several assumptions commonly made in AO imaging, assumptions which may be invalidated by the very high power and dioptric thickness of the rodent retina. We used optical modeling to compare the performance of rat and human eyes under conditions that tested the validity of these assumptions. Results showed that AO image quality in the human eye is robust to positioning errors of the AO corrector and to differences in imaging depth and wavelength compared to the wavefront beacon. In contrast, image quality in the rat eye declines sharply with each of these manipulations, especially when imaging off-axis. However, some latitude does exist to offset these manipulations against each other to produce good image quality. PMID:22876346

  6. Influence of affective image content on subjective quality assessment.

    PubMed

    van der Linde, Ian; Doe, Rachel M

    2012-09-01

    Image quality assessment (IQA) enables distortions introduced into an image (e.g., through lossy compression or broadcast) to be measured and evaluated for severity. It is unclear to what degree affective image content may influence this process. In this study, participants (n=25) were found to be unable to disentangle affective image content from objective image quality in a standard IQA procedure (single stimulus numerical categorical scale). We propose that this issue is worthy of consideration, particularly in single stimulus IQA techniques, in which a small number of handpicked images, not necessarily representative of the gamut of affect seen in true broadcasting, and unrated for affective content, serve as stimuli. PMID:23201952

  7. Automated FMV image quality assessment based on power spectrum statistics

    NASA Astrophysics Data System (ADS)

    Kalukin, Andrew

    2015-05-01

    Factors that degrade image quality in video and other sensor collections, such as noise, blurring, and poor resolution, also affect the spatial power spectrum of imagery. Prior research in human vision and image science from the last few decades has shown that the image power spectrum can be useful for assessing the quality of static images. The research in this article explores the possibility of using the image power spectrum to automatically evaluate full-motion video (FMV) imagery frame by frame. This procedure makes it possible to identify anomalous images and scene changes, and to keep track of gradual changes in quality as collection progresses. This article will describe a method to apply power spectral image quality metrics for images subjected to simulated blurring, blocking, and noise. As a preliminary test on videos from multiple sources, image quality measurements for image frames from 185 videos are compared to analyst ratings based on ground sampling distance. The goal of the research is to develop an automated system for tracking image quality during real-time collection, and to assign ratings to video clips for long-term storage, calibrated to standards such as the National Imagery Interpretability Rating System (NIIRS).

  8. Scatter radiation intensities around full-field digital mammography units

    PubMed Central

    Judge, M A; Keavey, E; Phelan, N

    2013-01-01

    ABSTRACT The aim of this study was to investigate the scatter radiation intensity around digital mammography systems and apply these data to standard shielding calculations to reveal whether shielding design of existing breast screening rooms is adequate for the use of digital mammography systems. Three digital mammography systems from GE Healthcare, Hologic and Philips were employed in the study. A breast-equivalent phantom was imaged under clinical workload conditions and scatter radiation intensities around the digital mammography systems were measured for a range of angles in three planes using an ionisation chamber. The results were compared with those from previous studies of film-screen systems. It may be deduced from the results that scattering in the backward direction is significant for all three systems, while scattering in the forward direction can be significant for some planes around the GE and Hologic systems. Measurements at typical clinical settings on each system revealed the Philips system to have markedly lower scatter radiation intensities than the other systems. Substituting the measured scattered radiation intensity into shielding calculations yielded barrier requirements similar to those already in place at the screening centres operating these systems. Current radiation protection requirements based on film-screen technology remain sufficient when applied to rooms with digital mammography installations and no alteration is required to the structural shielding. PMID:23239693

  9. High quality underwater imaging platform with laser range gated technique combining with image denoising and restoration

    NASA Astrophysics Data System (ADS)

    Huang, Huachuan; Wang, Rongbo; Yan, Keding; Yan, Zhengang; Wang, Shouyu; Li, Zhenhua; Li, Zeren

    2014-10-01

    Underwater laser imaging is of great significance in underwater search and marine science, etc. However, traditional underwater laser imaging is often of poor quality with noises and blurs, moreover, the resolution of the image is also low. In order to obtain clear underwater images with high resolution and quality, here, we have designed a range gated imaging underwater imaging system and realized an image restoration approach. In this paper, based on the introduction to the imaging system and image restoration algorithm, the experiment is established by setting the imaging system under water in the lake to capture the underwater targets. With the proposed underwater image restoration approach, images of high quality could be retrieved which proves that the method is able to identify the target ~10 meters away underwater.

  10. Issues to Consider in Converting to Digital Mammography

    PubMed Central

    Pisano, Etta D.; Zuley, Margarita; Baum, Janet K.; Marques, Helga S.

    2007-01-01

    This paper will outline the reasons that many radiology practices are converting to digital mammography. In addition, we will provide basic information on the issues that must be considered in making the transformation. These include technical matters regarding image display, storage and retrieval, as well as clinical and ergonomic considerations. PMID:17888771

  11. Mammographic equipment, technique, and quality control

    SciTech Connect

    Friedrich, M.A. )

    1991-08-01

    The most important improvements in mammographic technique were the introduction of single- or double-emulsion high-contrast film-screen combinations for mammography, the use of a specially designed low-kilovoltage Bucky grid to reduce scattered radiation, and the introduction of smaller focal spots to improve imaging geometry. Magnification techniques, especially the spot-film technique, yields clearer delineation of high-contrast microcalcifications. Dedicated mammographic equipment with specially designed x-ray tubes is necessary for modern high-quality mammography. However, in many modern mammographic units, the automatic exposure controller still fails to provide appropriate and constant optical film density over a wide range of tissue thickness and absorption. Extended-cycle processing of single-emulsion mammographic films can yield better image contrast and reduce exposure by up to 30%. Exposure times of less than 1 second are recommended to avoid the unnecessary higher doses caused by longer exposure times and reciprocity law failure. The wide dynamic range in mammography can be reduced by a beam equalization filter, and thus be better adapted to the decreased latitude of modern high-contrast mammographic screen-film systems. Mammographic film reading (detection of subtle microcalcifications) can be facilitated by modern computer evaluation of previously digitized mammograms. Standardization and assurance of image quality have been major challenges in the technical development of mammography. Different technical and anthropomorphic phantoms have been designed to measure and compare practical image quality. Detailed quality control measures have been developed. The benefit of a single or annual screening mammography, calculated in gained life expectancy, by far outweighs the relative risk for radiation-induced breast cancer. 22 references.

  12. Can pictorial images communicate the quality of pain successfully?

    PubMed

    Closs, S José; Knapp, Peter; Morley, Stephen; Stones, Catherine

    2015-08-01

    Chronic pain is common and difficult for patients to communicate to health professionals. It may include neuropathic elements which require specialised treatment. A little used approach to communicating the quality of pain is through the use of images. This study aimed to test the ability of a set of 12 images depicting different sensory pain qualities to successfully communicate those qualities. Images were presented to 25 student nurses and 38 design students. Students were asked to write down words or phrases describing the quality of pain they felt was being communicated by each image. They were asked to provide as many or as few as occurred to them. The images were extremely heterogeneous in their ability to convey qualities of pain accurately. Only 2 of the 12 images were correctly interpreted by more than 70% of the sample. There was a significant difference between the two student groups, with nurses being significantly better at interpreting the images than the design students. Clearly, attention needs to be given not only to the content of images designed to depict the sensory qualities of pain but also to the differing audiences who may use them. Education, verbal ability, ethnicity and a multiplicity of other factors may influence the understanding and use of such images. Considerable work is needed to develop a set of images which is sufficiently culturally appropriate and effective for general use. PMID:26516574

  13. Sensitometric analyses of screen-film systems for mammography exams in Brazil

    NASA Astrophysics Data System (ADS)

    Magalhães, L. A. G.; Drexler, G. G.; de Almeida, C. E.; Medeiros, L. L.; Ferreira, N. M. P. D.; Estrada, J. J. S.

    2015-12-01

    A determination of the sensitometric parameters of screen-film systems to evaluate their qualities was performed. The quality control of the automatic film processor was carried out to ensure a high level of efficiency. Based on ISO 9236-3, the following potentials were applied on the X-ray tubes: 25 kV, 28 kV, 30 kV and 35 kV. Four different mammography films from different manufacturers with and without screens were tested for curve shape, speed and average gradient. The results indicated that film 1 exhibited better contrast, film 3 demonstrated the highest energy dependence, and film 4 presented the largest base+fog density. None of the four mammographic films tested achieved satisfactory results in all parameters analyzed. Improvements in the manufacturing process for these films must be completed to avoid losses in the image quality.

  14. Dynamic flat panel detector versus image intensifier in cardiac imaging: dose and image quality

    NASA Astrophysics Data System (ADS)

    Vano, E.; Geiger, B.; Schreiner, A.; Back, C.; Beissel, J.

    2005-12-01

    The practical aspects of the dosimetric and imaging performance of a digital x-ray system for cardiology procedures were evaluated. The system was configured with an image intensifier (II) and later upgraded to a dynamic flat panel detector (FD). Entrance surface air kerma (ESAK) to phantoms of 16, 20, 24 and 28 cm of polymethyl methacrylate (PMMA) and the image quality of a test object were measured. Images were evaluated directly on the monitor and with numerical methods (noise and signal-to-noise ratio). Information contained in the DICOM header for dosimetry audit purposes was also tested. ESAK values per frame (or kerma rate) for the most commonly used cine and fluoroscopy modes for different PMMA thicknesses and for field sizes of 17 and 23 cm for II, and 20 and 25 cm for FD, produced similar results in the evaluated system with both technologies, ranging between 19 and 589 µGy/frame (cine) and 5 and 95 mGy min-1 (fluoroscopy). Image quality for these dose settings was better for the FD version. The 'study dosimetric report' is comprehensive, and its numerical content is sufficiently accurate. There is potential in the future to set those systems with dynamic FD to lower doses than are possible in the current II versions, especially for digital cine runs, or to benefit from improved image quality.

  15. Improving the Quality of Imaging in the Emergency Department.

    PubMed

    Blackmore, C Craig; Castro, Alexandra

    2015-12-01

    Imaging is critical for the care of emergency department (ED) patients. However, much of the imaging performed for acute care today is overutilization, creating substantial cost without significant benefit. Further, the value of imaging is not easily defined, as imaging only affects outcomes indirectly, through interaction with treatment. Improving the quality, including appropriateness, of emergency imaging requires understanding of how imaging contributes to patient care. The six-tier efficacy hierarchy of Fryback and Thornbury enables understanding of the value of imaging on multiple levels, ranging from technical efficacy to medical decision-making and higher-level patient and societal outcomes. The imaging efficacy hierarchy also allows definition of imaging quality through the Institute of Medicine (IOM)'s quality domains of safety, effectiveness, patient-centeredness, timeliness, efficiency, and equitability and provides a foundation for quality improvement. In this article, the authors elucidate the Fryback and Thornbury framework to define the value of imaging in the ED and to relate emergency imaging to the IOM quality domains. PMID:26568040

  16. Quaternion structural similarity: a new quality index for color images.

    PubMed

    Kolaman, Amir; Yadid-Pecht, Orly

    2012-04-01

    One of the most important issues for researchers developing image processing algorithms is image quality. Methodical quality evaluation, by showing images to several human observers, is slow, expensive, and highly subjective. On the other hand, a visual quality matrix (VQM) is a fast, cheap, and objective tool for evaluating image quality. Although most VQMs are good in predicting the quality of an image degraded by a single degradation, they poorly perform for a combination of two degradations. An example for such degradation is the color crosstalk (CTK) effect, which introduces blur with desaturation. CTK is expected to become a bigger issue in image quality as the industry moves toward smaller sensors. In this paper, we will develop a VQM that will be able to better evaluate the quality of an image degraded by a combined blur/desaturation degradation and perform as well as other VQMs on single degradations such as blur, compression, and noise. We show why standard scalar techniques are insufficient to measure a combined blur/desaturation degradation and explain why a vectorial approach is better suited. We introduce quaternion image processing (QIP), which is a true vectorial approach and has many uses in the fields of physics and engineering. Our new VQM is a vectorial expansion of structure similarity using QIP, which gave it its name-Quaternion Structural SIMilarity (QSSIM). We built a new database of a combined blur/desaturation degradation and conducted a quality survey with human subjects. An extensive comparison between QSSIM and other VQMs on several image quality databases-including our new database-shows the superiority of this new approach in predicting visual quality of color images. PMID:22203713

  17. Image processing system performance prediction and product quality evaluation

    NASA Technical Reports Server (NTRS)

    Stein, E. K.; Hammill, H. B. (Principal Investigator)

    1976-01-01

    The author has identified the following significant results. A new technique for image processing system performance prediction and product quality evaluation was developed. It was entirely objective, quantitative, and general, and should prove useful in system design and quality control. The technique and its application to determination of quality control procedures for the Earth Resources Technology Satellite NASA Data Processing Facility are described.

  18. Fractal analysis for reduced reference image quality assessment.

    PubMed

    Xu, Yong; Liu, Delei; Quan, Yuhui; Le Callet, Patrick

    2015-07-01

    In this paper, multifractal analysis is adapted to reduced-reference image quality assessment (RR-IQA). A novel RR-QA approach is proposed, which measures the difference of spatial arrangement between the reference image and the distorted image in terms of spatial regularity measured by fractal dimension. An image is first expressed in Log-Gabor domain. Then, fractal dimensions are computed on each Log-Gabor subband and concatenated as a feature vector. Finally, the extracted features are pooled as the quality score of the distorted image using l1 distance. Compared with existing approaches, the proposed method measures image quality from the perspective of the spatial distribution of image patterns. The proposed method was evaluated on seven public benchmark data sets. Experimental results have demonstrated the excellent performance of the proposed method in comparison with state-of-the-art approaches. PMID:25794391

  19. Optimization of contrast-enhanced spectral mammography depending on clinical indication.

    PubMed

    Dromain, Clarisse; Canale, Sandra; Saab-Puong, Sylvie; Carton, Ann-Katherine; Muller, Serge; Fallenberg, Eva Maria

    2014-10-01

    The objective is to optimize low-energy (LE) and high-energy (HE) exposure parameters of contrast-enhanced spectral mammography (CESM) examinations in four different clinical applications for which different levels of average glandular dose (AGD) and ratios between LE and total doses are required. The optimization was performed on a Senographe DS with a SenoBright® upgrade. Simulations were performed to find the optima by maximizing the contrast-to-noise ratio (CNR) on the recombined CESM image using different targeted doses and LE image quality. The linearity between iodine concentration and CNR as well as the minimal detectable iodine concentration was assessed. The image quality of the LE image was assessed on the CDMAM contrast-detail phantom. Experiments confirmed the optima found on simulation. The CNR was higher for each clinical indication than for SenoBright®, including the screening indication for which the total AGD was 22% lower. Minimal iodine concentrations detectable in the case of a 3-mm-diameter round tumor were 12.5% lower than those obtained for the same dose in the clinical routine. LE image quality satisfied EUREF acceptable limits for threshold contrast. This newly optimized set of acquisition parameters allows increased contrast detectability compared to parameters currently used without a significant loss in LE image quality. PMID:26158058

  20. Optimization of contrast-enhanced spectral mammography depending on clinical indication

    PubMed Central

    Dromain, Clarisse; Canale, Sandra; Saab-Puong, Sylvie; Carton, Ann-Katherine; Muller, Serge; Fallenberg, Eva Maria

    2014-01-01

    Abstract. The objective is to optimize low-energy (LE) and high-energy (HE) exposure parameters of contrast-enhanced spectral mammography (CESM) examinations in four different clinical applications for which different levels of average glandular dose (AGD) and ratios between LE and total doses are required. The optimization was performed on a Senographe DS with a SenoBright® upgrade. Simulations were performed to find the optima by maximizing the contrast-to-noise ratio (CNR) on the recombined CESM image using different targeted doses and LE image quality. The linearity between iodine concentration and CNR as well as the minimal detectable iodine concentration was assessed. The image quality of the LE image was assessed on the CDMAM contrast-detail phantom. Experiments confirmed the optima found on simulation. The CNR was higher for each clinical indication than for SenoBright®, including the screening indication for which the total AGD was 22% lower. Minimal iodine concentrations detectable in the case of a 3-mm-diameter round tumor were 12.5% lower than those obtained for the same dose in the clinical routine. LE image quality satisfied EUREF acceptable limits for threshold contrast. This newly optimized set of acquisition parameters allows increased contrast detectability compared to parameters currently used without a significant loss in LE image quality. PMID:26158058

  1. A new assessment method for image fusion quality

    NASA Astrophysics Data System (ADS)

    Li, Liu; Jiang, Wanying; Li, Jing; Yuchi, Ming; Ding, Mingyue; Zhang, Xuming

    2013-03-01

    Image fusion quality assessment plays a critically important role in the field of medical imaging. To evaluate image fusion quality effectively, a lot of assessment methods have been proposed. Examples include mutual information (MI), root mean square error (RMSE), and universal image quality index (UIQI). These image fusion assessment methods could not reflect the human visual inspection effectively. To address this problem, we have proposed a novel image fusion assessment method which combines the nonsubsampled contourlet transform (NSCT) with the regional mutual information in this paper. In this proposed method, the source medical images are firstly decomposed into different levels by the NSCT. Then the maximum NSCT coefficients of the decomposed directional images at each level are obtained to compute the regional mutual information (RMI). Finally, multi-channel RMI is computed by the weighted sum of the obtained RMI values at the various levels of NSCT. The advantage of the proposed method lies in the fact that the NSCT can represent image information using multidirections and multi-scales and therefore it conforms to the multi-channel characteristic of human visual system, leading to its outstanding image assessment performance. The experimental results using CT and MRI images demonstrate that the proposed assessment method outperforms such assessment methods as MI and UIQI based measure in evaluating image fusion quality and it can provide consistent results with human visual assessment.

  2. Toward a standard reference database for computer-aided mammography

    NASA Astrophysics Data System (ADS)

    Oliveira, Júlia E. E.; Gueld, Mark O.; de A. Araújo, Arnaldo; Ott, Bastian; Deserno, Thomas M.

    2008-03-01

    Because of the lack of mammography databases with a large amount of codified images and identified characteristics like pathology, type of breast tissue, and abnormality, there is a problem for the development of robust systems for computer-aided diagnosis. Integrated to the Image Retrieval in Medical Applications (IRMA) project, we present an available mammography database developed from the union of: The Mammographic Image Analysis Society Digital Mammogram Database (MIAS), The Digital Database for Screening Mammography (DDSM), the Lawrence Livermore National Laboratory (LLNL), and routine images from the Rheinisch-Westfälische Technische Hochschule (RWTH) Aachen. Using the IRMA code, standardized coding of tissue type, tumor staging, and lesion description was developed according to the American College of Radiology (ACR) tissue codes and the ACR breast imaging reporting and data system (BI-RADS). The import was done automatically using scripts for image download, file format conversion, file name, web page and information file browsing. Disregarding the resolution, this resulted in a total of 10,509 reference images, and 6,767 images are associated with an IRMA contour information feature file. In accordance to the respective license agreements, the database will be made freely available for research purposes, and may be used for image based evaluation campaigns such as the Cross Language Evaluation Forum (CLEF). We have also shown that it can be extended easily with further cases imported from a picture archiving and communication system (PACS).

  3. Triple-energy contrast enhanced digital mammography

    NASA Astrophysics Data System (ADS)

    Puong, Sylvie; Milioni de Carvalho, Pablo; Muller, Serge

    2010-04-01

    With the injection of iodine, Contrast Enhanced Digital Mammography (CEDM) provides functional information about breast tumour angiogenesis that can potentially help in cancer diagnosis. In order to generate iodine images in which the gray level is proportional to the iodine thickness, temporal and dual-energy approaches have already been considered. The dual-energy method offers the advantage of less patient motion artifacts and better comfort during the exam. However, this approach requires knowledge of the breast thickness at each pixel. Generally, as compression is applied, the breast thickness at each pixel is taken as the compression thickness. Nevertheless, in the breast border region, this assumption is not correct anymore and this causes inaccuracies in the iodine image. Triple-Energy CEDM could overcome these limitations by providing supplemental information in the form of a third image acquired with a different spectrum than the other two. This precludes the need of a priori knowledge of the breast thickness. Moreover, with Triple-Energy CEDM, breast thickness and glandularity maps could potentially be derived. In this study, we first focused on the method to recombine the three images in order to generate the iodine image, analyzing the performance of either quadratic, cubic or conic recombination functions. Then, we studied the optimal acquisition spectra in order to maximize the iodine SDNR in the recombined image for a given target total glandular dose. The concept of Triple-Energy CEDM was validated on simulated textured images and poly-energetic images acquired with a conventional X-ray mammography tube.

  4. Comparison of signal to noise ratios from spatial and frequency domain formulations of nonprewhitening model observers in digital mammography

    SciTech Connect

    Sisini, Francesco; Zanca, Federica; Marshall, Nicholas W.; Taibi, Angelo; Cardarelli, Paolo; Bosmans, Hilde

    2012-09-15

    Purpose: Image quality indices based upon model observers are promising alternatives to laborious human readings of contrast-detail images. This is especially appealing in digital mammography as limiting values for contrast thresholds determine, according to some international protocols, the acceptability of these systems in the radiological practice. The objective of the present study was to compare the signal to noise ratios (SNR) obtained with two nonprewhitening matched filter model observer approaches, one in the spatial domain and the other in the frequency domain, and with both of them worked out for disks as present in the CDMAM phantom. Methods: The analysis was performed using images acquired with the Siemens Novation and Inspiration digital mammography systems. The spatial domain formulation uses a series of high dose CDMAM images as the signal and a routine exposure of two flood images to calculate the covariance matrix. The frequency domain approach uses the mathematical description of a disk and modulation transfer function (MTF) and noise power spectrum (NPS) calculated from images. Results: For both systems most of the SNR values calculated in the frequency domain were in very good agreement with the SNR values calculated in the spatial domain. Both the formulations in the frequency domain and in the spatial domain show a linear relationship between SNR and the diameter of the CDMAM discs. Conclusions: The results suggest that both formulations of the model observer lead to very similar figures of merit. This is a step forward in the adoption of figures of merit based on NPS and MTF for the acceptance testing of mammography systems.

  5. Comparison of the astronomical and multimedia image quality criteria

    NASA Astrophysics Data System (ADS)

    Anisimova, Elena; Páta, Petr; Fliegel, Karel; Klíma, Miloš

    2012-06-01

    This paper deals with the criteria definition of image quality in astronomy and their comparison with common multimedia approaches. Astronomical images have typical specific properties - high grayscale bit depth, size, high noise occurrence, sensitivity to point spread function deformation and special processing algorithms. They belong to the class of scientific images as well as medical or similar. Their processing and compression is quite different from the classical approach of multimedia image processing. The new compression algorithm based on JPEG2000 is selected as a distortion source in this paper. Selected image quality criteria (multimedia and optimized for astronomical images) are tested on the set of images from the DEIMOS image database with miscellaneous level of the thermally generated CCD noise. The deformation of the point spread function (PSF) is also measured for chosen compression approach.

  6. No-reference visual quality assessment for image inpainting

    NASA Astrophysics Data System (ADS)

    Voronin, V. V.; Frantc, V. A.; Marchuk, V. I.; Sherstobitov, A. I.; Egiazarian, K.

    2015-03-01

    Inpainting has received a lot of attention in recent years and quality assessment is an important task to evaluate different image reconstruction approaches. In many cases inpainting methods introduce a blur in sharp transitions in image and image contours in the recovery of large areas with missing pixels and often fail to recover curvy boundary edges. Quantitative metrics of inpainting results currently do not exist and researchers use human comparisons to evaluate their methodologies and techniques. Most objective quality assessment methods rely on a reference image, which is often not available in inpainting applications. Usually researchers use subjective quality assessment by human observers. It is difficult and time consuming procedure. This paper focuses on a machine learning approach for no-reference visual quality assessment for image inpainting based on the human visual property. Our method is based on observation that Local Binary Patterns well describe local structural information of the image. We use a support vector regression learned on assessed by human images to predict perceived quality of inpainted images. We demonstrate how our predicted quality value correlates with qualitative opinion in a human observer study. Results are shown on a human-scored dataset for different inpainting methods.

  7. Improvement of image quality by polarization mixing

    NASA Astrophysics Data System (ADS)

    Kasahara, Ryosuke; Itoh, Izumi; Hirai, Hideaki

    2014-03-01

    Information about the polarization of light is valuable because it contains information about the light source illuminating an object, the illumination angle, and the object material. However, polarization information strongly depends on the direction of the light source, and it is difficult to use a polarization image with various recognition algorithms outdoors because the angle of the sun varies. We propose an image enhancement method for utilizing polarization information in many such situations where the light source is not fixed. We take two approaches to overcome this problem. First, we compute an image that is the combination of a polarization image and the corresponding brightness image. Because of the angle of the light source, the polarization contains no information about some scenes. Therefore, it is difficult to use only polarization information in any scene for applications such as object detection. However, if we use a combination of a polarization image and a brightness image, the brightness image can complement the lack of scene information. The second approach is finding features that depend less on the direction of the light source. We propose a method for extracting scene features based on a calculation of the reflection model including polarization effects. A polarization camera that has micro-polarizers on each pixel of the image sensor was built and used for capturing images. We discuss examples that demonstrate the improved visibility of objects by applying our proposed method to, e.g., the visibility of lane markers on wet roads.

  8. Meat quality evaluation by hyperspectral imaging technique: an overview.

    PubMed

    Elmasry, Gamal; Barbin, Douglas F; Sun, Da-Wen; Allen, Paul

    2012-01-01

    During the last two decades, a number of methods have been developed to objectively measure meat quality attributes. Hyperspectral imaging technique as one of these methods has been regarded as a smart and promising analytical tool for analyses conducted in research and industries. Recently there has been a renewed interest in using hyperspectral imaging in quality evaluation of different food products. The main inducement for developing the hyperspectral imaging system is to integrate both spectroscopy and imaging techniques in one system to make direct identification of different components and their spatial distribution in the tested product. By combining spatial and spectral details together, hyperspectral imaging has proved to be a promising technology for objective meat quality evaluation. The literature presented in this paper clearly reveals that hyperspectral imaging approaches have a huge potential for gaining rapid information about the chemical structure and related physical properties of all types of meat. In addition to its ability for effectively quantifying and characterizing quality attributes of some important visual features of meat such as color, quality grade, marbling, maturity, and texture, it is able to measure multiple chemical constituents simultaneously without monotonous sample preparation. Although this technology has not yet been sufficiently exploited in meat process and quality assessment, its potential is promising. Developing a quality evaluation system based on hyperspectral imaging technology to assess the meat quality parameters and to ensure its authentication would bring economical benefits to the meat industry by increasing consumer confidence in the quality of the meat products. This paper provides a detailed overview of the recently developed approaches and latest research efforts exerted in hyperspectral imaging technology developed for evaluating the quality of different meat products and the possibility of its widespread deployment. PMID:22591341

  9. SAR image quality using advanced pulse compression noise (APCN)

    NASA Astrophysics Data System (ADS)

    Govoni, Mark A.; Elwell, Ryan A.

    2014-05-01

    This work demonstrates the feasibility of using the advanced pulse compression noise (APCN) radar waveform for synthetic aperture radar (SAR). Using a simple image formation process (IFP), we not only show that we can successfully form images using the APCN waveform, but we grow our understanding of how different combinations of APCN waveforms and side lobe weighting functions impact SAR image quality. In this paper, an analysis is presented that compares the target range point spread function (PSF) for several simulated SAR images.

  10. Univariant assessment of the visual quality of images

    NASA Astrophysics Data System (ADS)

    Jung, Mathieu; Leger, Dominique

    2000-06-01

    In order to evaluate the visual quality of images, most methods compare a degraded image to a perfect reference. We propose an original univariant (i.e. without reference) method based on the use of artificial neural networks. The principle is first to use a neural network to learn the quality of images taken from a pool of known examples, then use it to assess the quality of unknown images. The considered defects are compression artefacts, ringing or local singularities. To simplify, only images with defects that are not mixed with each other were first used. The method follows four steps. Observers are first required to mark degraded images to establish a pool of examples. Then, a characterization of the defect is extracted mathematically from the image. Then, the neural network learns how to establish a relation between the mathematical characterization of the defect and the visual mark. Finally, it can be used to assess the visual quality of an unknown image from the mathematical characterization of its defects. Two illustrative examples are presented: the assessment of the quality of JPEG compressed images and the detection of local defects.

  11. Quality evaluation of digital fundus images through combined measures.

    PubMed

    Veiga, Diana; Pereira, Carla; Ferreira, Manuel; Gonçalves, Luís; Monteiro, João

    2014-04-01

    The evaluation of image quality is an important step before an automatic analysis of retinal images. Several conditions can impair the acquisition of a good image, and minimum image quality requirements should be present to ensure that an automatic or semiautomatic system provides an accurate diagnosis. A method to classify fundus images as low or good quality is presented. The method starts with the detection of regions of uneven illumination and evaluates if the segmented noise masks affect a clinically relevant area (around the macula). Afterwards, focus is evaluated through a fuzzy classifier. An input vector is created extracting three focus features. The system was validated in a large dataset (1454 fundus images), obtained from an online database and an eye clinic and compared with the ratings of three observers. The system performance was close to optimal with an area under the receiver operating characteristic curve of 0.9943. PMID:26158021

  12. Evidence Base for Quality Control Activities in Cardiovascular Imaging.

    PubMed

    Eskandari, Mehdi; Kramer, Christopher M; Hecht, Harvey S; Jaber, Wael A; Marwick, Thomas H

    2016-03-01

    Quality control is pervasive in most modern business, but, surprisingly, is in its infancy in medicine in general-and cardiovascular imaging in particular. The increasing awareness of the cost of cardiovascular imaging, matched by a desire to show benefits from imaging to patient outcome, suggests that this deficiency should be reassessed. Demonstration of improved quality has been proposed to require a focus on several domains: laboratory organization, patient selection, image acquisition, image interpretation, and results communication. Improvement in these steps will require adoption of a variety of interventions, including laboratory accreditation, appropriate use criteria, and continuous quality control and enhancements in reporting, but the evidence base for the benefit of interventions on these steps has been sparse. The purpose of this review is to evaluate the current status and future goals of developing the evidence base for these processes in cardiovascular imaging. PMID:26965731

  13. Quality evaluation of digital fundus images through combined measures

    PubMed Central

    Veiga, Diana; Pereira, Carla; Ferreira, Manuel; Gonçalves, Luís; Monteiro, João

    2014-01-01

    Abstract. The evaluation of image quality is an important step before an automatic analysis of retinal images. Several conditions can impair the acquisition of a good image, and minimum image quality requirements should be present to ensure that an automatic or semiautomatic system provides an accurate diagnosis. A method to classify fundus images as low or good quality is presented. The method starts with the detection of regions of uneven illumination and evaluates if the segmented noise masks affect a clinically relevant area (around the macula). Afterwards, focus is evaluated through a fuzzy classifier. An input vector is created extracting three focus features. The system was validated in a large dataset (1454 fundus images), obtained from an online database and an eye clinic and compared with the ratings of three observers. The system performance was close to optimal with an area under the receiver operating characteristic curve of 0.9943. PMID:26158021

  14. Update on new technologies in digital mammography

    PubMed Central

    Patterson, Stephanie K; Roubidoux, Marilyn A

    2014-01-01

    Despite controversy regarding mammography’s efficacy, it continues to be the most commonly used breast cancer-screening modality. With the development of digital mammography, some improved benefit has been shown in women with dense breast tissue. However, the density of breast tissue continues to limit the sensitivity of conventional mammography. We discuss the development of some derivative digital technologies, primarily digital breast tomosynthesis, and their strengths, weaknesses, and potential patient impact. PMID:25152634

  15. Effect of acquisition parameters on image quality in digital tomosynthesis

    NASA Astrophysics Data System (ADS)

    Deller, Timothy; Jabri, Kadri N.; Sabol, John M.; Ni, Xianfeng; Avinash, Gopal; Saunders, Rowland; Uppaluri, Renuka

    2007-03-01

    Digital tomosynthesis (DTS) is emerging as an advanced imaging technique that enables volumetric slice imaging with a detector typically used for projection radiography. An understanding of the interactions between DTS acquisition parameters and characteristics of the reconstructed slice images is required for optimizing the acquisition protocols of various clinical applications. This paper presents our investigation of the effects and interactions of acquisition parameters, including sweep angle, number of projections, and dose, on clinically relevant image-quality metrics. Metrics included the image characteristics of in-slice resolution, depth resolution, image noise level, and presence of ripple. Phantom experiments were performed to characterize the relationship between the acquisition parameters and image quality. Results showed that the depth resolution was mainly dependent on sweep angle. Visibility of ripple was determined by the projection density (number of projections divided by sweep angle), as well as properties of the imaged object. Image noise was primarily dependent on total dose and not significantly affected by the number of projections. These experimental and theoretical results were confirmed using anthropomorphic phantoms and also used to develop clinical acquisition protocols. Assessment of phantom and clinical images obtained with these protocols revealed that the use of acquisition protocols optimized for a given clinical exam enables rapid, low-dose, high quality DTS imaging for diverse clinical applications including abdomen, hand, shoulder, spine, and chest. We conclude that DTS acquisition parameters have a significant effect on image quality and should be tailored for the imaged anatomy and desired clinical application. Relationships developed in this work will guide the selection of acquisition protocols to improve image quality and clinical utility of DTS for a wide variety of clinical exams.

  16. Image quality assessment by preprocessing and full reference model combination

    NASA Astrophysics Data System (ADS)

    Bianco, S.; Ciocca, G.; Marini, F.; Schettini, R.

    2009-01-01

    This paper focuses on full-reference image quality assessment and presents different computational strategies aimed to improve the robustness and accuracy of some well known and widely used state of the art models, namely the Structural Similarity approach (SSIM) by Wang and Bovik and the S-CIELAB spatial-color model by Zhang and Wandell. We investigate the hypothesis that combining error images with a visual attention model could allow a better fit of the psycho-visual data of the LIVE Image Quality assessment Database Release 2. We show that the proposed quality assessment metric better correlates with the experimental data.

  17. Characterization tests of a homemade ionization chamber in mammography standard radiation beams

    NASA Astrophysics Data System (ADS)

    Silva, J. O.; Nonato, F. B. C.; Caldas, L. V. E.

    2014-02-01

    A mammography homemade ionization chamber was developed to be applied for mammography energy range dosimetry. This chamber has a sensitive volume of 6 cm3 and is made of a Lucite body and graphite coated collecting electrode. Characteristics such as saturation, ion collection efficiency, linearity of chamber response versus air kerma rate and energy dependence were determined. The results obtained with the mammography homemade ionization chamber are within the limits stated in international recommendations. This chamber can be used in quality control programs in the diagnostic radiology area. All measurements were carried out at the Calibration Laboratory of IPEN.

  18. Quantification of breast arterial calcification using full field digital mammography

    SciTech Connect

    Molloi, Sabee; Xu Tong; Ducote, Justin; Iribarren, Carlos

    2008-04-15

    Breast arterial calcification is commonly detected on some mammograms. Previous studies indicate that breast arterial calcification is evidence of general atherosclerotic vascular disease and it may be a useful marker of coronary artery disease. It can potentially be a useful tool for assessment of coronary artery disease in women since mammography is widely used as a screening tool for early detection of breast cancer. However, there are currently no available techniques for quantification of calcium mass using mammography. The purpose of this study was to determine whether it is possible to quantify breast arterial calcium mass using standard digital mammography. An anthropomorphic breast phantom along with a vessel calcification phantom was imaged using a full field digital mammography system. Densitometry was used to quantify calcium mass. A calcium calibration measurement was performed at each phantom thickness and beam energy. The known (K) and measured (M) calcium mass on 5 and 9 cm thickness phantoms were related by M=0.964K-0.288 mg (r=0.997 and SEE=0.878 mg) and M=1.004K+0.324 mg (r=0.994 and SEE=1.32 mg), respectively. The results indicate that accurate calcium mass measurements can be made without correction for scatter glare as long as careful calcium calibration is made for each breast thickness. The results also indicate that composition variations and differences of approximately 1 cm between calibration phantom and breast thickness introduce only minimal error in calcium measurement. The uncertainty in magnification is expected to cause up to 5% and 15% error in calcium mass for 5 and 9 cm breast thicknesses, respectively. In conclusion, a densitometry technique for quantification of breast arterial calcium mass was validated using standard full field digital mammography. The results demonstrated the feasibility and potential utility of the densitometry technique for accurate quantification of breast arterial calcium mass using standard digital mammography.

  19. Quantification of breast arterial calcification using full field digital mammography

    PubMed Central

    Molloi, Sabee; Xu, Tong; Ducote, Justin; Iribarren, Carlos

    2008-01-01

    Breast arterial calcification is commonly detected on some mammograms. Previous studies indicate that breast arterial calcification is evidence of general atherosclerotic vascular disease and it may be a useful marker of coronary artery disease. It can potentially be a useful tool for assessment of coronary artery disease in women since mammography is widely used as a screening tool for early detection of breast cancer. However, there are currently no available techniques for quantification of calcium mass using mammography. The purpose of this study was to determine whether it is possible to quantify breast arterial calcium mass using standard digital mammography. An anthropomorphic breast phantom along with a vessel calcification phantom was imaged using a full field digital mammography system. Densitometry was used to quantify calcium mass. A calcium calibration measurement was performed at each phantom thickness and beam energy. The known (K) and measured (M) calcium mass on 5 and 9 cm thickness phantoms were related by M=0.964K?0.288 mg (r=0.997 and SEE=0.878 mg) and M=1.004K+0.324 mg (r=0.994 and SEE=1.32 mg), respectively. The results indicate that accurate calcium mass measurements can be made without correction for scatter glare as long as careful calcium calibration is made for each breast thickness. The results also indicate that composition variations and differences of approximately 1 cm between calibration phantom and breast thickness introduce only minimal error in calcium measurement. The uncertainty in magnification is expected to cause up to 5% and 15% error in calcium mass for 5 and 9 cm breast thicknesses, respectively. In conclusion, a densitometry technique for quantification of breast arterial calcium mass was validated using standard full field digital mammography. The results demonstrated the feasibility and potential utility of the densitometry technique for accurate quantification of breast arterial calcium mass using standard digital mammography. PMID:18491538

  20. Fundamental limits of positron emission mammography

    SciTech Connect

    Moses, William W.; Qi, Jinyi

    2001-06-01

    We explore the causes of performance limitation in positron emission mammography cameras. We compare two basic camera geometries containing the same volume of 511 keV photon detectors, one with a parallel plane geometry and another with a rectangular geometry. We find that both geometries have similar performance for the phantom imaged (in Monte Carlo simulation), even though the solid angle coverage of the rectangular camera is about 50 percent higher than the parallel plane camera. The reconstruction algorithm used significantly affects the resulting image; iterative methods significantly outperform the commonly used focal plane tomography. Finally, the characteristics of the tumor itself, specifically the absolute amount of radiotracer taken up by the tumor, will significantly affect the imaging performance.

  1. Barriers to Mammography among Inadequately Screened Women

    ERIC Educational Resources Information Center

    Stoll, Carolyn R. T.; Roberts, Summer; Cheng, Meng-Ru; Crayton, Eloise V.; Jackson, Sherrill; Politi, Mary C.

    2015-01-01

    Mammography use has increased over the past 20 years, yet more than 30% of women remain inadequately screened. Structural barriers can deter individuals from screening, however, cognitive, emotional, and communication barriers may also prevent mammography use. This study sought to identify the impact of number and type of barriers on mammography…

  2. Image quality assessment using Takagi-Sugeno-Kang fuzzy model

    NASA Astrophysics Data System (ADS)

    Ä?ordevi?, Dragana; Kukolj, Dragan; Schelkens, Peter

    2015-03-01

    The main aim of the paper is to present a non-linear image quality assessment model based on a fuzzy logic estimator, namely the Takagi-Sugeno-Kang fuzzy model. This image quality assessment model uses a clustered space of input objective metrics. Main advantages of the introduced quality model are simplicity and understandably of its fuzzy rules. As reference model the polynomial 3 rd order model was chosen. The parameters of the Takagi-Sugeno-Kang fuzzy model are optimized in accordance to the mapping criteria of the selected set of input objective quality measures to the Mean Opinion Score (MOS) scale.

  3. Digital Mammography with a Mosaic of CCD Arrays

    NASA Technical Reports Server (NTRS)

    Jalink, Antony, Jr. (Inventor); McAdoo, James A. (Inventor)

    1998-01-01

    A digital mammography device uses a mosaic of electronic digital imaging arrays to scan an x-ray image is discussed. The mosaic of arrays is repositioned several times to expose different portions of the image, until the entire image is scanned. The data generated by the arrays during each exposure is stored in a computer. After the final exposure, the computer combines data of the several partial images to produce a composite of the original x-ray image. An aperture plate is used to reduce scatter and the overall exposure of the patient to x-rays.

  4. Feature maps driven no-reference image quality prediction of authentically distorted images

    NASA Astrophysics Data System (ADS)

    Ghadiyaram, Deepti; Bovik, Alan C.

    2015-03-01

    Current blind image quality prediction models rely on benchmark databases comprised of singly and synthetically distorted images, thereby learning image features that are only adequate to predict human perceived visual quality on such inauthentic distortions. However, real world images often contain complex mixtures of multiple distortions. Rather than a) discounting the effect of these mixtures of distortions on an image's perceptual quality and considering only the dominant distortion or b) using features that are only proven to be efficient for singly distorted images, we deeply study the natural scene statistics of authentically distorted images, in different color spaces and transform domains. We propose a feature-maps-driven statistical approach which avoids any latent assumptions about the type of distortion(s) contained in an image, and focuses instead on modeling the remarkable consistencies in the scene statistics of real world images in the absence of distortions. We design a deep belief network that takes model-based statistical image features derived from a very large database of authentically distorted images as input and discovers good feature representations by generalizing over different distortion types, mixtures, and severities, which are later used to learn a regressor for quality prediction. We demonstrate the remarkable competence of our features for improving automatic perceptual quality prediction on a benchmark database and on the newly designed LIVE Authentic Image Quality Challenge Database and show that our approach of combining robust statistical features and the deep belief network dramatically outperforms the state-of-the-art.

  5. Photoreceptor waveguides and effective retinal image quality

    NASA Astrophysics Data System (ADS)

    Vohnsen, Brian

    2007-03-01

    Individual photoreceptor waveguiding suggests that the entire retina can be considered as a composite fiber-optic element relating a retinal image to a corresponding waveguided image. In such a scheme, a visual sensation is produced only when the latter interacts with the pigments of the outer photoreceptor segments. Here the possible consequences of photoreceptor waveguiding on vision are studied with important implications for the pupil-apodization method commonly used to incorporate directional effects of the retina. In the absence of aberrations, it is found that the two approaches give identical predictions for an effective retinal image only when the pupil apodization is chosen twice as narrow as suggested by the traditional Stiles-Crawford effect. In addition, phase variations in the retinal field due to ocular aberrations can delicately alter a waveguided image, and this may provide plausible justification for an improved visual sensation as compared with what should be expected on the grounds of a retinal image only.

  6. Objective analysis of image quality of video image capture systems

    NASA Astrophysics Data System (ADS)

    Rowberg, Alan H.

    1990-07-01

    As Picture Archiving and Communication System (PACS) technology has matured, video image capture has become a common way of capturing digital images from many modalities. While digital interfaces, such as those which use the ACR/NEMA standard, will become more common in the future, and are preferred because of the accuracy of image transfer, video image capture will be the dominant method in the short term, and may continue to be used for some time because of the low cost and high speed often associated with such devices. Currently, virtually all installed systems use methods of digitizing the video signal that is produced for display on the scanner viewing console itself. A series of digital test images have been developed for display on either a GE CT9800 or a GE Signa MRI scanner. These images have been captured with each of five commercially available image capture systems, and the resultant images digitally transferred on floppy disk to a PC1286 computer containing Optimast' image analysis software. Here the images can be displayed in a comparative manner for visual evaluation, in addition to being analyzed statistically. Each of the images have been designed to support certain tests, including noise, accuracy, linearity, gray scale range, stability, slew rate, and pixel alignment. These image capture systems vary widely in these characteristics, in addition to the presence or absence of other artifacts, such as shading and moire pattern. Other accessories such as video distribution amplifiers and noise filters can also add or modify artifacts seen in the captured images, often giving unusual results. Each image is described, together with the tests which were performed using them. One image contains alternating black and white lines, each one pixel wide, after equilibration strips ten pixels wide. While some systems have a slew rate fast enough to track this correctly, others blur it to an average shade of gray, and do not resolve the lines, or give horizontal or vertical streaking. While many of these results are significant from an engineering standpoint alone, there are clinical implications and some anatomy or pathology may not be visualized if an image capture system is used improperly.

  7. 78 FR 13681 - Agency Information Collection Activities; Proposed Collection; Comment Request; Mammography...

    Federal Register 2010, 2011, 2012, 2013, 2014

    2013-02-28

    ..., personnel, and practices, including quality assurance. The intent of these regulations is to assure safe... showing that they meet the equipment, personnel, quality assurance and quality control standards, and have... Collection; Comment Request; Mammography Quality Standards Act Requirements AGENCY: Food and...

  8. Slot scanning versus antiscatter grid in digital mammography: comparison of low-contrast performance using contrast-detail measurement

    NASA Astrophysics Data System (ADS)

    Lai, Chao-Jen; Shaw, Chris C.; Geiser, William; Kappadath, Srinivas C.; Liu, Xinming; Wang, TianPeng; Tu, Shu-Ju; Altunbas, Mustafa C.

    2004-05-01

    Slot scanning imaging techniques allow for effective scatter rejection without attenuating primary x-rays. The use of these techniques should generate better image quality for the same mean glandular dose (MGD) or a similar image quality for a lower MGD as compared to imaging techniques using an anti-scatter grid. In this study, we compared a slot scanning digital mammography system (SenoScan, Fisher Imaging Systems, Denver, CO) to a full-field digital mammography (FFDM) system used in conjunction with a 5:1 anti-scatter grid (SenoGraphe 2000D, General Electric Medical Systems, Milwaukee, WI). Images of a contrast-detail phantom (University Hospital Nijmegen, The Netherlands) were reviewed to measure the contrast-detail curves for both systems. These curves were measured at 100%, 71%, 49% and 33% of the reference mean glandular dose (MGD), as determined by photo-timing, for the Fisher system and 100% for the GE system. Soft-copy reading was performed on review workstations provided by the manufacturers. The correct observation ratios (CORs) were also computed and used to compare the performance of the two systems. The results showed that, based on the contrast-detail curves, the performance of the Fisher images, acquired at 100% and 71% of the reference MGD, was comparable to the GE images at 100% of the reference MGD. The CORs for Fisher images were 0.463 and 0.444 at 100% and 71% of the reference MGD, respectively, compared to 0.453 for the GE images at 100% of the reference MGD.

  9. Digital Receptor Image Quality Evaluation: Effect of Different Filtration Schemes

    NASA Astrophysics Data System (ADS)

    Murphy, Simon; Christianson, Olav; Amurao, Maxwell; Samei, Ehsan

    2010-04-01

    The International Electrotechnical Commission provides a standard measurement methodology to provide performance intercomparison between imaging systems. Its formalism specifies beam quality based on half value layer attained by target kVp and additional Al filtration. Similar beam quality may be attained more conveniently using a filtration combination of Cu and Al. This study aimed to compare the two filtration schemes by their effects on image quality in terms of signal-difference-to-noise ratio, spatial resolution, exposure index, noise power spectrum, modulation transfer function, and detective quantum efficiency. A comparative assessment of the images was performed by analyzing commercially available image quality assessment phantom and by following the IEC 62220-3 formalism.

  10. The clinical effectiveness and cost-effectiveness of different surveillance mammography regimens after the treatment for primary breast cancer: systematic reviews registry database analyses and economic evaluation.

    PubMed Central

    Robertson, C; Arcot Ragupathy, S K; Boachie, C; Dixon, J M; Fraser, C; Hernández, R; Heys, S; Jack, W; Kerr, G R; Lawrence, G; MacLennan, G; Maxwell, A; McGregor, J; Mowatt, G; Pinder, S; Ternent, L; Thomas, R E; Vale, L; Wilson, R; Zhu, S; Gilbert, F J

    2011-01-01

    BACKGROUND Following primary breast cancer treatment, the early detection of ipsilateral breast tumour recurrence (IBTR) or ipsilateral secondary cancer in the treated breast and detection of new primary cancers in the contralateral breast is beneficial for survival. Surveillance mammography is used to detect these cancers, but the optimal frequency of surveillance and the length of follow-up are unclear. OBJECTIVES To identify feasible management strategies for surveillance and follow-up of women after treatment for primary breast cancer in a UK setting, and to determine the effectiveness and cost-effectiveness of differing regimens. METHODS A survey of UK breast surgeons and radiologists to identify current surveillance mammography regimens and inform feasible alternatives; two discrete systematic reviews of evidence published from 1990 to mid 2009 to determine (i) the clinical effectiveness and cost-effectiveness of differing surveillance mammography regimens for patient health outcomes and (ii) the test performance of surveillance mammography in the detection of IBTR and metachronous contralateral breast cancer (MCBC); statistical analysis of individual patient data (West Midlands Cancer Intelligence Unit Breast Cancer Registry and Edinburgh data sets); and economic modelling using the systematic reviews results, existing data sets, and focused searches for specific data analysis to determine the effectiveness and cost-utility of differing surveillance regimens. RESULTS The majority of survey respondents initiate surveillance mammography 12 months after breast-conserving surgery (BCS) (87%) or mastectomy (79%). Annual surveillance mammography was most commonly reported for women after BCS or after mastectomy (72% and 53%, respectively). Most (74%) discharge women from surveillance mammography, most frequently 10 years after surgery. The majority (82%) discharge from clinical follow-up, most frequently at 5 years. Combining initiation, frequency and duration of surveillance mammography resulted in 54 differing surveillance regimens for women after BCS and 56 for women following mastectomy. The eight studies included in the clinical effectiveness systematic review suggest surveillance mammography offers a survival benefit compared with a surveillance regimen that does not include surveillance mammography. Nine studies were included in the test performance systematic review. For routine IBTR detection, surveillance mammography sensitivity ranged from 64% to 67% and specificity ranged from 85% to 97%. For magnetic resonance imaging (MRI), sensitivity ranged from 86% to 100% and specificity was 93%. For non-routine IBTR detection, sensitivity and specificity for surveillance mammography ranged from 50% to 83% and from 57% to 75%, respectively, and for MRI from 93% to 100% and from 88% to 96%, respectively. For routine MCBC detection, one study reported sensitivity of 67% and specificity of 50% for both surveillance mammography and MRI, although this was a highly select population. Data set analysis showed that IBTR has an adverse effect on survival. Furthermore, women experiencing a second tumour measuring >20 mm in diameter were at a significantly greater risk of death than those with no recurrence or those whose tumour was <10 mm in diameter. In the base-case analysis, the strategy with the highest net benefit, and most likely to be considered cost-effective, was surveillance mammography alone, provided every 12 months at a societal willingness to pay for a quality-adjusted life-year of either £20,000 or £30,000. The incremental cost-effectiveness ratio for surveillance mammography alone every 12 months compared with no surveillance was £4727. LIMITATIONS Few studies met the review inclusion criteria and none of the studies was a randomised controlled trial. The limited and variable nature of the data available precluded any quantitative analysis. There was no useable evidence contained in the Breast Cancer Registry database to assess the effectiveness of surveillance mammography directly. The results of the economic model should be considered exploratory and interpreted with caution given the paucity of data available to inform the economic model. CONCLUSIONS Surveillance is likely to improve survival and patients should gain maximum benefit through optimal use of resources, with those women with a greater likelihood of developing IBTR or MCBC being offered more comprehensive and more frequent surveillance. Further evidence is required to make a robust and informed judgement on the effectiveness of surveillance mammography and follow-up. The utility of national data sets could be improved and there is a need for high-quality, direct head-to-head studies comparing the diagnostic accuracy of tests used in the surveillance population. FUNDING The National Institute for Health Research Health Technology Assessment programme. PMID:21951942

  11. A single-photon counting “edge-on” silicon detector for synchrotron radiation mammography

    NASA Astrophysics Data System (ADS)

    Rigon, L.; Arfelli, F.; Astolfo, A.; Bergamaschi, A.; Dreossi, D.; Longo, R.; Menk, R.-H.; Schmitt, B.; Vallazza, E.; Castelli, E.

    2009-09-01

    The Phase Imaging for Clinical Application with Silicon detector and Synchrotron radiatiOn (PICASSO) project is developing an "edge-on" silicon microstrip detector for mammography with synchrotron radiation. The sensor is equipped with a fast single-photon counting electronics based on the Mythen-II application-specific integrated circuit. A first prototype has been assembled and tested at the SYnchrotron Radiation for MEdical Physics (SYRMEP) beamline at Elettra in Trieste, Italy. The first results are presented in this study including evidence of high-rate single-photon counting with negligible losses up to 1.2×10 6 incident photons per pixel per second; spatial resolution consistent with the pixel aperture (0.3 mm×0.05 mm); high-quality imaging of test-objects, obtained with a dose comparable to the one delivered in modern full-field digital mammographic systems.

  12. Dosimetry and image quality assessment in a direct radiography system

    PubMed Central

    Oliveira, Bruno Beraldo; de Oliveira, Marcio Alves; Paixão, Lucas; Teixeira, Maria Helena Araújo; Nogueira, Maria do Socorro

    2014-01-01

    Objective To evaluate the mean glandular dose with a solid state detector and the image quality in a direct radiography system, utilizing phantoms. Materials and Methods Irradiations were performed with automatic exposure control and polymethyl methacrylate slabs with different thicknesses to calculate glandular dose values. The image quality was evaluated by means of the structures visualized on the images of the phantoms. Results Considering the uncertainty of the measurements, the mean glandular dose results are in agreement with the values provided by the equipment and with internationally adopted reference levels. Results obtained from images of the phantoms were in agreement with the reference values. Conclusion The present study contributes to verify the equipment conformity as regards dose values and image quality. PMID:25741119

  13. Evaluation of lossy data compression in primary interpretation for full-field digital mammography.

    PubMed

    Kovacs, Mark D; Reicher, Joshua J; Grotts, Jonathan F; Reicher, Murray A; Trambert, Michael A

    2015-03-01

    OBJECTIVE. For full-field digital mammography (FFDM), federal regulations prohibit lossy data compression for primary reading and archiving, unlike all other medical images, where reading physicians can apply their professional judgment in implementing lossy compression. Faster image transfer, lower costs, and greater access to expert mammographers would result from development of a safe standard for primary interpretation and archive of lossy-compressed FFDM images. This investigation explores whether JPEG 2000 80:1 lossy data compression affects clinical accuracy in digital mammography. MATERIALS AND METHODS. Randomized FFDM cases (n = 194) were interpreted by six experienced mammographers with and without JPEG 2000 80:1 lossy compression applied. A cancer-enriched population was used, with just less than half of the cases (42%) containing subtle (< 1 cm) biopsy-proven cancerous lesions, and the remaining cases were negative as proven by 2-year follow-up. Data were analyzed using the jackknife alternative free-response ROC (JAFROC) method. RESULTS. The differences in reader performance between lossy-compressed and non-lossy-compressed images using lesion localization (0.660 vs 0.671), true-positive fraction (0.879 vs 0.879), and false-positive fraction (0.283 vs 0.271) were not statistically significant. There was no difference in the JAFROC figure of merit between lossy-compressed and non-lossy-compressed images, with a mean difference of -0.01 (95% CI, -0.03 to 0.01; F1,5 = 2.30; p = 0.189). CONCLUSION. These results suggest that primary interpretation of JPEG 2000 80:1 lossy-compressed FFDM images may be viable without degradation of clinical quality. Benefits would include lower storage costs, faster telemammography, and enhanced access to expert mammographers. PMID:25714287

  14. The use of the general image quality equation in the design and evaluation of imaging systems

    NASA Astrophysics Data System (ADS)

    Cota, Steve A.; Florio, Christopher J.; Duvall, David J.; Leon, Michael A.

    2009-08-01

    The design of any modern imaging system is the end result of many trade studies, each seeking to optimize image quality within real world constraints such as cost, schedule and overall risk. The National Imagery Interpretability Rating Scale (NIIRS) is a useful measure of image quality, because, by characterizing the overall interpretability of an image, it combines into one metric those contributors to image quality to which a human interpreter is most sensitive. The main drawback to using a NIIRS rating as a measure of image quality in engineering trade studies is the fact that it is tied to the human observer and cannot be predicted from physical principles and engineering parameters alone. The General Image Quality Equation (GIQE) of Leachtenauer et al. 1997 [Appl. Opt. 36, 8322-8328 (1997)] is a regression of actual image analyst NIIRS ratings vs. readily calculable engineering metrics, and provides a mechanism for using the expected NIIRS rating of an imaging system in the design and evaluation process. In this paper, we will discuss how we use the GIQE in conjunction with The Aerospace Corporation's Parameterized Image Chain Analysis & Simulation SOftware (PICASSO) to evaluate imager designs, taking a hypothetical high resolution commercial imaging system as an example.

  15. Is the supply of mammography machines outstripping need and demand? An economic analysis.

    PubMed

    Brown, M L; Kessler, L G; Rueter, F G

    1990-10-01

    The number of dedicated mammography machines installed in the United States has grown explosively. It is estimated that almost 10,000 machines will be installed by 1990, whereas the projected demand for screening mammography will require only approximately 2,600 machines, if the machines are used in a moderately efficient manner. The excess supply of mammography resources raises concern from an economic perspective for several reasons. First, such a condition means that health care resources are being used inefficiently. Second, the low average utilization rate of mammography equipment implied by these results necessitates charging a high price-over $100, on average-to cover costs. This price is above the $50 usually associated with low-cost screening mammography programs, and it may impede a desirable public health trend to increase use of mammography screening. Third, the existence of many mammography facilities operating at low capacity levels is inefficient from a health systems perspective, increasing the cost of quality assurance and medical record keeping. The current condition of excess supply is probably unsustainable over the long term. PMID:2393209

  16. Raman chemical imaging technology for food safety and quality evaluation

    Technology Transfer Automated Retrieval System (TEKTRAN)

    Raman chemical imaging combines Raman spectroscopy and digital imaging to visualize composition and morphology of a target. This technique offers great potential for food safety and quality research. Most commercial Raman instruments perform measurement at microscopic level, and the spatial range ca...

  17. Objective quality assessment of tone-mapped images.

    PubMed

    Yeganeh, Hojatollah; Wang, Zhou

    2013-02-01

    Tone-mapping operators (TMOs) that convert high dynamic range (HDR) to low dynamic range (LDR) images provide practically useful tools for the visualization of HDR images on standard LDR displays. Different TMOs create different tone-mapped images, and a natural question is which one has the best quality. Without an appropriate quality measure, different TMOs cannot be compared, and further improvement is directionless. Subjective rating may be a reliable evaluation method, but it is expensive and time consuming, and more importantly, is difficult to be embedded into optimization frameworks. Here we propose an objective quality assessment algorithm for tone-mapped images by combining: 1) a multiscale signal fidelity measure on the basis of a modified structural similarity index and 2) a naturalness measure on the basis of intensity statistics of natural images. Validations using independent subject-rated image databases show good correlations between subjective ranking score and the proposed tone-mapped image quality index (TMQI). Furthermore, we demonstrate the extended applications of TMQI using two examples-parameter tuning for TMOs and adaptive fusion of multiple tone-mapped images. PMID:23047872

  18. A feature-enriched completely blind image quality evaluator.

    PubMed

    Lin Zhang; Lei Zhang; Bovik, Alan C

    2015-08-01

    Existing blind image quality assessment (BIQA) methods are mostly opinion-aware. They learn regression models from training images with associated human subjective scores to predict the perceptual quality of test images. Such opinion-aware methods, however, require a large amount of training samples with associated human subjective scores and of a variety of distortion types. The BIQA models learned by opinion-aware methods often have weak generalization capability, hereby limiting their usability in practice. By comparison, opinion-unaware methods do not need human subjective scores for training, and thus have greater potential for good generalization capability. Unfortunately, thus far no opinion-unaware BIQA method has shown consistently better quality prediction accuracy than the opinion-aware methods. Here, we aim to develop an opinion-unaware BIQA method that can compete with, and perhaps outperform, the existing opinion-aware methods. By integrating the features of natural image statistics derived from multiple cues, we learn a multivariate Gaussian model of image patches from a collection of pristine natural images. Using the learned multivariate Gaussian model, a Bhattacharyya-like distance is used to measure the quality of each image patch, and then an overall quality score is obtained by average pooling. The proposed BIQA method does not need any distorted sample images nor subjective quality scores for training, yet extensive experiments demonstrate its superior quality-prediction performance to the state-of-the-art opinion-aware BIQA methods. The MATLAB source code of our algorithm is publicly available at www.comp.polyu.edu.hk/~cslzhang/IQA/ILNIQE/ILNIQE.htm. PMID:25915960

  19. Contrast-Enhanced Digital Mammography and Angiogenesis

    SciTech Connect

    Rosado-Mendez, I.; Palma, B. A.; Villasenor, Y.; Benitez-Bribiesca, L.; Brandan, M. E.

    2007-11-26

    Angiogenesis could be a means for pouring contrast media around tumors. In this work, optimization of radiological parameters for contrast-enhanced subtraction techniques in mammography has been performed. A modification of Lemacks' analytical formalism was implemented to model the X-ray absorption in the breast with contrast medium and detection by a digital image receptor. Preliminary results of signal-to-noise ratio analysis show the advantage of subtracting two images taken at different energies, one prior and one posterior to the injection of contrast medium. Preliminary experimental results using a custom-made phantom have shown good agreement with calculations. A proposal is presented for the clinical application of the optimized technique, which aims at finding correlations between angiogenesis indicators and dynamic variables of contrast medium uptake.

  20. WE-A-12A-01: Medical Physics 1.0 to 2.0, Session 2: Radiography, Mammography and Fluoroscopy

    SciTech Connect

    Gingold, E; Karellas, A; Strauss, K

    2014-06-15

    Medical Physics 2.0 is a bold vision for an existential transition of clinical imaging physics in face of the new realities of value-based and evidencebased medicine, comparative effectiveness, and meaningful use. It speaks to how clinical imaging physics can expand beyond traditional insular models of inspection and acceptance testing, oriented toward compliance, towards team-based models of operational engagement, prospective definition and assurance of effective use, and retrospective evaluation of clinical performance. Organized into four sessions of the AAPM, this particular session focuses on three specific modalities as outlined below. Radiography 2.0: The development of electronic capture in recent years has changed the landscape and spurred reinvestment by healthcare providers. The radiography presentation will explore how the diagnostic medical physicist must adapt to these changes to support radiographic imaging, and how she/he can add value in radiography practice over the next 5-10 years. Topics of discussion include new metrology of evaluation, new models of clinical engagement, and effective integration of new technologies. Mammography 2.0: Mammography has been an interesting testing ground on the effectiveness of close involvement of medical physicists with equipment in the past twenty years. The outcomes have clearly shown major improvements in image quality and significant reduction in the average glandular dose. However, the medical physicist's role in mammography has been largely focused to annual surveys and with limited input on operational issues with image artifacts, optimal mammographic acquisition mode and problems with image quality. This mammography presentation will address why and how medical physicists must be prepared to address the new models of practice that include new metrics of performance and the integration of new technologies (DBT, syncretized mammograms, contrast mammography, breast CT) into clinical practice. Fluoroscopy 2.0: Physics support of fluoroscopy should be operationally as opposed to compliance focused. Testing protocols must address new hardware, acquisition methods, and image processing. Future available tools are discussed. Proper configuration of acquisition parameters (focal spot size, voltage and added filter, tube current, pulse width, pulse rate, scatter removal) as a function of patient size from the neonate to bariatric patient is key to providing diagnostic image quality at properly managed radiation doses. Learning Objectives: Appreciate the limitations of the currently available tools and techniques in clinical medical physics in radiography, mammography, and fluoroscopy, and ways to improve upon current deficiencies. Appreciate the changing environment of imaging practice and the need for the medical physicist to be an expert consultant and educator in a capacity that extends beyond the annual survey of equipment. Understand the status of the rapidly changing environment in breast imaging from planar imaging to tomosynthesis and possibly to breast CT. Identify appropriate configuration of acquisition parameters as a function of patient size to manage radiation dose and ensure diagnostic image quality.

  1. Determination of Tube Output (kVp) and Exposure Mode for Breast Phantom of Various Thicknesses/Glandularity for Digital Mammography

    PubMed Central

    IZDIHAR, Kamal; KANAGA, Kumari Chelliah; KRISHNAPILLAI, Vijayalakshimi; SULAIMAN, Tamanang

    2015-01-01

    Background: Optimisation of average glandular dose (AGD) for two-dimensional (2D) mammography is important, as imaging using ionizing radiation has the probability to induce cancer resulting from stochastic effects. This study aims to observe the effects of kVp, anode/filter material, and exposure mode on the dose and image quality of 2D mammography. Methods: This experimental study was conducted using full-field digital mammography. The entrance surface air kerma was determined using thermoluminescent dosimeter (TLD) 100H and ionization chamber (IC) on three types of Computerized Imaging Reference System (CIRS) phantom with 50/50, 30/70, and 20/80 breast glandularity, respectively, in the auto-time mode and auto-filter mode. The Euref protocol was used to calculate the AGD while the image quality was evaluated using contrast-to-noise ratio (CNR), figure of merit (FOM), and image quality figure (IQF). Results: It is shown that AGD values in the auto-time mode did not decrease significantly with the increasing tube voltage of the silver filter (r = −0.187, P > 0.05) and rhodium filter (r = −0.131, P > 0.05) for all the phantoms. The general linear model showed that AGD for all phantoms had a significant effect between different exposure factors [F (6,12.3) = 4.48 and mode of exposure F (1,86) = 4.17, P < 0.05, respectively] but there is no significant difference between the different anode/filter combination [F (1,4) = 0.571]. Conclusion: In summary, the 28, 29, and 31 kVp are the optimum kVp for 50%, 30%, and 20% breast glandularity, respectively. Besides the auto-filter mode is suitable for 50%, 30%, and 20% breast glandularity because it is automatic, faster, and may avoid error done by the operator. PMID:25892949

  2. Perceived quality of wood images influenced by the skewness of image histogram

    NASA Astrophysics Data System (ADS)

    Katsura, Shigehito; Mizokami, Yoko; Yaguchi, Hirohisa

    2015-08-01

    The shape of image luminance histograms is related to material perception. We investigated how the luminance histogram contributed to improvements in the perceived quality of wood images by examining various natural wood and adhesive vinyl sheets with printed wood grain. In the first experiment, we visually evaluated the perceived quality of wood samples. In addition, we measured the colorimetric parameters of the wood samples and calculated statistics of image luminance. The relationship between visual evaluation scores and image statistics suggested that skewness and kurtosis affected the perceived quality of wood. In the second experiment, we evaluated the perceived quality of wood images with altered luminance skewness and kurtosis using a paired comparison method. Our result suggests that wood images are more realistic if the skewness of the luminance histogram is slightly negative.

  3. Measuring Satisfaction with Mammography Results Reporting

    PubMed Central

    Dolan, Nancy C; Feinglass, Joe; Priyanath, Aparna; Haviley, Corrine; Sorensen, Asta V; Venta, Luz A

    2001-01-01

    OBJECTIVE To assess factors associated with patient satisfaction with communication of mammography results and their understanding and ability to recall these results. DESIGN Cross-sectional telephone survey. SETTING Academic breast imaging center. PATIENTS Two hundred ninety-eight patients who had either a screening or diagnostic mammogram. MEASUREMENTS AND MAIN RESULTS Survey items assessed waiting time for results, anxiety about results, satisfaction with several components of results reporting, and patients' understanding of results and recommendations. Women undergoing screening exams were more likely to be dissatisfied with the way the results were communicated than those who underwent diagnostic exams and received immediate results (20% vs 11%, P = .05). For these screening patients, waiting for more than two weeks for notification of results, difficulty getting in touch with someone to answer questions, low ratings of how clearly results were explained, and considerable or extreme anxiety about the results were all independently associated with dissatisfaction with the way the results were reported, while age and actual exam result were not. CONCLUSIONS Patients undergoing screening mammograms were more likely to be dissatisfied with the way the results were communicated than were those who underwent diagnostic mammograms. Interventions to reduce the wait time for results, reduce patients' anxiety, and improve the clarity with which the results and recommendations are given may help improve overall satisfaction with mammography result reporting. PMID:11318910

  4. Development of a fast read-out system of a single photon counting detector for mammography with synchrotron radiation

    NASA Astrophysics Data System (ADS)

    Lopez, F. C.; Rigon, L.; Longo, R.; Arfelli, F.; Bergamaschi, A.; Chen, R. C.; Dreossi, D.; Schmitt, B.; Vallazza, E.; Castelli, E.

    2011-12-01

    A single-photon counting detector read-out system for mammography with synchrotron radiation has been developed with the aim to meet the needs of the mammographic imaging station of the SYRMEP beamline at ELETTRA. The system called PICASSO (Phase Imaging for Clinical Application with Silicon detector and Synchrotron radiatiOn) is a modular detector that implements a read-out system with MYTHEN II ASICs, an embedded Linux-based controller board and a Scientific Linux acquisition workstation. The system architecture and characteristics are herein presented. The system was tested at the SYRMEP beamline and achieved a frame rate of 33 Hz for 8448 channels at 24-bit dynamic range, and it is capable of continuously acquiring up to 2000 frames. Standard mammographic phantoms were imaged and good quality images were obtained at doses comparable with what is delivered in conventional full field mammographic systems.

  5. Digital image quality measurements by objective and subjective methods from series of parametrically degraded images

    NASA Astrophysics Data System (ADS)

    Tachó, Aura; Mitjà, Carles; Martínez, Bea; Escofet, Jaume; Ralló, Miquel

    2013-11-01

    Many digital image applications like digitization of cultural heritage for preservation purposes operate with compressed files in one or more image observing steps. For this kind of applications JPEG compression is one of the most widely used. Compression level, final file size and quality loss are parameters that must be managed optimally. Although this loss can be monitored by means of objective image quality measurements, the real challenge is to know how it can be related with the perceived image quality by observers. A pictorial image has been degraded by two different procedures. The first, applying different levels of low pass filtering by convolving the image with progressively broad Gauss kernels. The second, saving the original file to a series of JPEG compression levels. In both cases, the objective image quality measurement is done by analysis of the image power spectrum. In order to obtain a measure of the perceived image quality, both series of degraded images are displayed on a computer screen organized in random pairs. The observers are compelled to choose the best image of each pair. Finally, a ranking is established applying Thurstone scaling method. Results obtained by both measurements are compared between them and with other objective measurement method as the Slanted Edge Test.

  6. Pre-analytic process control: projecting a quality image.

    PubMed

    Serafin, Mark D

    2006-01-01

    Within the health-care system, the term "ancillary department" often describes the laboratory. Thus, laboratories may find it difficult to define their image and with it, customer perception of department quality. Regulatory requirements give laboratories who so desire an elegant way to address image and perception issues--a comprehensive pre-analytic system solution. Since large laboratories use such systems--laboratory service manuals--I describe and illustrate the process for the benefit of smaller facilities. There exist resources to help even small laboratories produce a professional service manual--an elegant solution to image and customer perception of quality. PMID:17005095

  7. Effect of optical aberrations on image quality and visual performance

    NASA Astrophysics Data System (ADS)

    Ravikumar, Sowmya

    In addition to the effects of diffraction, retinal image quality in the human eye is degraded by optical aberrations. Although the paraxial geometric optics description of defocus consists of a simple blurred circle whose size determines the extent of blur, in reality the interactions between monochromatic and chromatic aberrations create a complex pattern of retinal image degradation. My thesis work hypothesizes that although both monochromatic and chromatic optical aberrations in general reduce image quality from best achievable, the underlying causes of retinal image quality degradation are characteristic of the nature of the aberration, its interactions with other aberrations as well as the composition of the stimulus. To establish a controlled methodology, a computational model of the retinal image with various levels of aberrations was used to create filters equivalent to those produced by real optical aberrations. Visual performance was measured psychophysically by using these special filters that separately modulated amplitude and phase in the retinal image. In order to include chromatic aberration into the optical interactions, a computational polychromatic model of the eye was created and validated. The model starts with monochromatic wavefront maps and derives a composite white light point-spread function whose quality was assessed using metrics of image quality. Finally, in order to assess the effectiveness of simultaneous multifocal intra-ocular lenses in correcting the eye's optical aberrations, a polychromatic computational model of a pseudophakic eye was constructed. This model incorporated the special chromatic properties unique to an eye corrected with hybrid refractive-diffractive optical elements. Results showed that normal optical aberrations reduced visual performance not only by reducing image contrast but also by altering the phase structure of the image. Longitudinal chromatic aberration had a greater effect on image quality in isolation than in the presence of monochromatic aberrations. Also, the diffractive optical element was found to improve polychromatic image quality in a pseudophakic eye by chromatic correction. My thesis work shows that in order to obtain maximal improvement in image quality, it is important to correct both monochromatic and chromatic aberrations.

  8. Theoretical analysis of high-resolution digital mammography.

    PubMed

    Suryanarayanan, Sankararaman; Karellas, Andrew; Vedantham, Srinivasan; Sechopoulos, Ioannis

    2006-06-21

    The performance of a high-resolution charge coupled device-based full-field digital mammography imager was analysed using a mathematical framework based on an adaptation of cascaded linear systems theory described by other investigators. This work has been conducted in order to understand the impact of various design parameters on the physical performance characteristics of the imager. Specifically, the effect of pixel size, scintillator thickness and packing density, x-ray spectra, air kerma, dark current, charge integration time, and pixel fill-factor on the frequency dependent detective quantum efficiency was studied using a charge-coupled device as a reference platform. The imaging system was modelled as a series of physical processes with gain and spatial spreading. For each stage, the signal and noise power spectra were computed and propagated through the imaging chain as inputs to subsequent stages. Good agreement between experimental and theoretical predictions was obtained for various x-ray spectral conditions that were investigated. The modulation transfer function, MTF(f) and detective quantum efficiency DQE(f) characteristics obtained in this study are encouraging and comparable to other digital mammography systems. The results of this study strongly suggest the feasibility of large area scintillator-based digital mammography imagers with pixel sizes below 100 microm. PMID:16757861

  9. Optimization and image quality assessment of the alpha-image reconstruction algorithm: iterative reconstruction with well-defined image quality metrics

    NASA Astrophysics Data System (ADS)

    Lebedev, Sergej; Sawall, Stefan; Kuchenbecker, Stefan; Faby, Sebastian; Knaup, Michael; Kachelrieß, Marc

    2015-03-01

    The reconstruction of CT images with low noise and highest spatial resolution is a challenging task. Usually, a trade-off between at least these two demands has to be found or several reconstructions with mutually exclusive properties, i.e. either low noise or high spatial resolution, have to be performed. Iterative reconstruction methods might be suitable tools to overcome these limitations and provide images of highest diagnostic quality with formerly mutually exclusive image properties. While image quality metrics like the modulation transfer function (MTF) or the point spread function (PSF) are well-defined in case of standard reconstructions, e.g. filtered backprojection, the iterative algorithms lack these metrics. To overcome this issue alternate methodologies like the model observers have been proposed recently to allow a quantification of a usually task-dependent image quality metric.1 As an alternative we recently proposed an iterative reconstruction method, the alpha-image reconstruction (AIR), providing well-defined image quality metrics on a per-voxel basis.2 In particular, the AIR algorithm seeks to find weighting images, the alpha-images, that are used to blend between basis images with mutually exclusive image properties. The result is an image with highest diagnostic quality that provides a high spatial resolution and a low noise level. As the estimation of the alpha-images is computationally demanding we herein aim at optimizing this process and highlight the favorable properties of AIR using patient measurements.

  10. Study of a water quality imager for coastal zone missions

    NASA Technical Reports Server (NTRS)

    Staylor, W. F.; Harrison, E. F.; Wessel, V. W.

    1975-01-01

    The present work surveys water quality user requirements and then determines the general characteristics of an orbiting imager (the Applications Explorer, or AE) dedicated to the measurement of water quality, which could be used as a low-cost means of testing advanced imager concepts and assessing the ability of imager techniques to meet the goals of a comprehensive water quality monitoring program. The proposed imager has four spectral bands, a spatial resolution of 25 meters, and swath width of 36 km with a pointing capability of 330 km. Silicon photodetector arrays, pointing systems, and several optical features are included. A nominal orbit of 500 km altitude at an inclination of 50 deg is recommended.

  11. Implications of Overdiagnosis: Impact on Screening Mammography Practices

    PubMed Central

    Morris, Elizabeth; Feig, Stephen A.; Drexler, Madeline

    2015-01-01

    Abstract This review article explores the issue of overdiagnosis in screening mammography. Overdiagnosis is the screen detection of a breast cancer, histologically confirmed, that might not otherwise become clinically apparent during the lifetime of the patient. While screening mammography is an imperfect tool, it remains the best tool we have to diagnose breast cancer early, before a patient is symptomatic and at a time when chances of survival and options for treatment are most favorable. In 2015, an estimated 231,840 new cases of breast cancer (excluding ductal carcinoma in situ) will be diagnosed in the United States, and some 40,290 women will die. Despite these data, screening mammography for women ages 40–69 has contributed to a substantial reduction in breast cancer mortality, and organized screening programs have led to a shift from late-stage diagnosis to early-stage detection. Current estimates of overdiagnosis in screening mammography vary widely, from 0% to upwards of 30% of diagnosed cancers. This range reflects the fact that measuring overdiagnosis is not a straightforward calculation, but usually one based on different sets of assumptions and often biased by methodological flaws. The recent development of tomosynthesis, which creates high-resolution, three-dimensional images, has increased breast cancer detection while reducing false recalls. Because the greatest harm of overdiagnosis is overtreatment, the key goal should not be less diagnosis but better treatment decision tools. (Population Health Management 2015;18:S3–S11) PMID:26414384

  12. Noisy images-JPEG compressed: subjective and objective image quality evaluation

    NASA Astrophysics Data System (ADS)

    Corchs, Silvia; Gasparini, Francesca; Schettini, Raimondo

    2014-01-01

    The aim of this work is to study image quality of both single and multiply distorted images. We address the case of images corrupted by Gaussian noise or JPEG compressed as single distortion cases and images corrupted by Gaussian noise and then JPEG compressed, as multiply distortion case. Subjective studies were conducted in two parts to obtain human judgments on the single and multiply distorted images. We study how these subjective data correlate with No Reference state-of-the-art quality metrics. We also investigate proper combining of No Reference metrics to achieve better performance. Results are analyzed and compared in terms of correlation coefficients.

  13. Objective Quality Assessment for Color-to-Gray Image Conversion.

    PubMed

    Ma, Kede; Zhao, Tiesong; Zeng, Kai; Wang, Zhou

    2015-12-01

    Color-to-gray (C2G) image conversion is the process of transforming a color image into a grayscale one. Despite its wide usage in real-world applications, little work has been dedicated to compare the performance of C2G conversion algorithms. Subjective evaluation is reliable but is also inconvenient and time consuming. Here, we make one of the first attempts to develop an objective quality model that automatically predicts the perceived quality of C2G converted images. Inspired by the philosophy of the structural similarity index, we propose a C2G structural similarity (C2G-SSIM) index, which evaluates the luminance, contrast, and structure similarities between the reference color image and the C2G converted image. The three components are then combined depending on image type to yield an overall quality measure. Experimental results show that the proposed C2G-SSIM index has close agreement with subjective rankings and significantly outperforms existing objective quality metrics for C2G conversion. To explore the potentials of C2G-SSIM, we further demonstrate its use in two applications: 1) automatic parameter tuning for C2G conversion algorithms and 2) adaptive fusion of C2G converted images. PMID:26208349

  14. The effect of image quality and forensic expertise in facial image comparisons.

    PubMed

    Norell, Kristin; Läthén, Klas Brorsson; Bergström, Peter; Rice, Allyson; Natu, Vaidehi; O'Toole, Alice

    2015-03-01

    Images of perpetrators in surveillance video footage are often used as evidence in court. In this study, identification accuracy was compared for forensic experts and untrained persons in facial image comparisons as well as the impact of image quality. Participants viewed thirty image pairs and were asked to rate the level of support garnered from their observations for concluding whether or not the two images showed the same person. Forensic experts reached their conclusions with significantly fewer errors than did untrained participants. They were also better than novices at determining when two high-quality images depicted the same person. Notably, lower image quality led to more careful conclusions by experts, but not for untrained participants. In summary, the untrained participants had more false negatives and false positives than experts, which in the latter case could lead to a higher risk of an innocent person being convicted for an untrained witness. PMID:25537273

  15. Peripheral Aberrations and Image Quality for Contact Lens Correction

    PubMed Central

    Shen, Jie; Thibos, Larry N.

    2011-01-01

    Purpose Contact lenses reduced the degree of hyperopic field curvature present in myopic eyes and rigid contact lenses reduced sphero-cylindrical image blur on the peripheral retina, but their effect on higher order aberrations and overall optical quality of the eye in the peripheral visual field is still unknown. The purpose of our study was to evaluate peripheral wavefront aberrations and image quality across the visual field before and after contact lens correction. Methods A commercial Hartmann-Shack aberrometer was used to measure ocular wavefront errors in 5° steps out to 30° of eccentricity along the horizontal meridian in uncorrected eyes and when the same eyes are corrected with soft or rigid contact lenses. Wavefront aberrations and image quality were determined for the full elliptical pupil encountered in off-axis measurements. Results Ocular higher-order aberrations increase away from fovea in the uncorrected eye. Third-order aberrations are larger and increase faster with eccentricity compared to the other higher-order aberrations. Contact lenses increase all higher-order aberrations except 3rd-order Zernike terms. Nevertheless, a net increase in image quality across the horizontal visual field for objects located at the foveal far point is achieved with rigid lenses, whereas soft contact lenses reduce image quality. Conclusions Second order aberrations limit image quality more than higher-order aberrations in the periphery. Although second-order aberrations are reduced by contact lenses, the resulting gain in image quality is partially offset by increased amounts of higher-order aberrations. To fully realize the benefits of correcting higher-order aberrations in the peripheral field requires improved correction of second-order aberrations as well. PMID:21873925

  16. Learning Receptive Fields and Quality Lookups for Blind Quality Assessment of Stereoscopic Images.

    PubMed

    Shao, Feng; Lin, Weisi; Wang, Shanshan; Jiang, Gangyi; Yu, Mei; Dai, Qionghai

    2016-03-01

    Blind quality assessment of 3D images encounters more new challenges than its 2D counterparts. In this paper, we propose a blind quality assessment for stereoscopic images by learning the characteristics of receptive fields (RFs) from perspective of dictionary learning, and constructing quality lookups to replace human opinion scores without performance loss. The important feature of the proposed method is that we do not need a large set of samples of distorted stereoscopic images and the corresponding human opinion scores to learn a regression model. To be more specific, in the training phase, we learn local RFs (LRFs) and global RFs (GRFs) from the reference and distorted stereoscopic images, respectively, and construct their corresponding local quality lookups (LQLs) and global quality lookups (GQLs). In the testing phase, blind quality pooling can be easily achieved by searching optimal GRF and LRF indexes from the learnt LQLs and GQLs, and the quality score is obtained by combining the LRF and GRF indexes together. Experimental results on three publicly 3D image quality assessment databases demonstrate that in comparison with the existing methods, the devised algorithm achieves high consistent alignment with subjective assessment. PMID:25872220

  17. A Dynamic Image Quality Evaluation of Videofluoroscopy Images: Considerations for Telepractice Applications.

    PubMed

    Burns, Clare L; Keir, Benjamin; Ward, Elizabeth C; Hill, Anne J; Farrell, Anna; Phillips, Nick; Porter, Linda

    2015-08-01

    High-quality fluoroscopy images are required for accurate interpretation of videofluoroscopic swallow studies (VFSS) by speech pathologists and radiologists. Consequently, integral to developing any system to conduct VFSS remotely via telepractice is ensuring that the quality of the VFSS images transferred via the telepractice system is optimized. This study evaluates the extent of change observed in image quality when videofluoroscopic images are transmitted from a digital fluoroscopy system to (a) current clinical equipment (KayPentax Digital Swallowing Workstation, and b) four different telepractice system configurations. The telepractice system configurations consisted of either a local C20 or C60 Cisco TelePresence System (codec unit) connected to the digital fluoroscopy system and linked to a second remote C20 or C60 Cisco TelePresence System via a network running at speeds of either 2, 4 or 6 megabits per second (Mbit/s). Image quality was tested using the NEMA XR 21 Phantom, and results demonstrated some loss in spatial resolution, low contrast detectability and temporal resolution for all transferred images when compared to the fluoroscopy source. When using higher capacity codec units and/or the highest bandwidths to support data transmission, image quality transmitted through the telepractice system was found to be comparable if not better than the current clinical system. This study confirms that telepractice systems can be designed to support fluoroscopy image transfer and highlights important considerations when developing telepractice systems for VFSS analysis to ensure high-quality radiological image reproduction. PMID:26014137

  18. Image science and image-quality research in the Optical Sciences Center

    NASA Astrophysics Data System (ADS)

    Barrett, Harrison H.; Myers, Kyle J.

    2014-09-01

    This paper reviews the history of research into imaging and image quality at the Optical Sciences Center (OSC), with emphasis on the period 1970-1990. The work of various students in the areas of psychophysical studies of human observers of images; mathematical model observers; image simulation and analysis, and the application of these methods to radiology and nuclear medicine is summarized. The rapid progress in computational power, at OSC and elsewhere, which enabled the steady advances in imaging and the emergence of a science of imaging, is also traced. The implications of these advances to ongoing research and the current Image Science curriculum at the College of Optical Sciences are discussed.

  19. Multiple-image encryption based on triple interferences for flexibly decrypting high-quality images.

    PubMed

    Li, Wei-Na; Phan, Anh-Hoang; Piao, Mei-Lan; Kim, Nam

    2015-04-10

    We propose a multiple-image encryption (MIE) scheme based on triple interferences for flexibly decrypting high-quality images. Each image is discretionarily deciphered without decrypting a series of other images earlier. Since it does not involve any cascaded encryption orders, the image can be decrypted flexibly by using the novel method. Computer simulation demonstrated that the proposed method's running time is less than approximately 1/4 that of the previous similar MIE method. Moreover, the decrypted image is perfectly correlated with the original image, and due to many phase functions serving as decryption keys, this method is more secure and robust. PMID:25967313

  20. Experience with a proposed teleradiology system for digital mammography

    NASA Astrophysics Data System (ADS)

    Saulnier, Emilie T.; Mitchell, Robert J.; Abdel-Malek, Aiman A.; Dudding, Kathryn E.

    1995-05-01

    Teleradiology offers significant improvement in efficiency and effectiveness over current practices in traditional film/screen-based diagnosis. In the context of digital mammography, the increasing number of women who need to be screened for breast cancer, including those in remote rural regions, make the advantages of teleradiology especially attractive for digital mammography. At the same time, the size and resolution of digital mammograms are among the most challenging to support in a cost effective teleradiology system. This paper describes a teleradiology architecture developed for use with digital mammography by GE Corporate Research and Development in collaboration with Massachusetts General Hospital under National Cancer Institute (NCI/NIH) grant number R01 CA60246-01. Experience with a testbed prototype is described. The telemammography architecture is intended to consist of a main mammography diagnostic site serving several remote screening sites. As patient exams become available, they are forwarded by an image server to the diagnostic site over a WAN communications link. A radiologist at the diagnostic site views a patient exam as it arrives, interprets it, and then relays a report back to the technician at the remote site. A secondary future scenario consists of mobile units which forward images to a remote site, which then forwards them to the main diagnostic site. The testbed architecture is based on the Digital Imaging and Communications in Medicine (DICOM) standard, created by the American College of Radiology (ACR) and National Electrical Manufacturers Association (NEMA). A specification of vendor-independent data formats and data transfer services for digital medical images, DICOM specifies a protocol suite starting at the application layer downward, including the TCP/IP layers. The current DICOM definition does not provide an information element that is specifically tailored to mammography, so we have used the DICOM secondary capture data format for the mammography images. In conclusion, experience with the testbed is described, as is performance analysis related to selection of network components needed to extend this architecture to clinical evaluation. Recommendations are made as to the critical areas for future work.

  1. APQ-102 imaging radar digital image quality study

    NASA Technical Reports Server (NTRS)

    Griffin, C. R.; Estes, J. M.

    1982-01-01

    A modified APQ-102 sidelooking radar collected synthetic aperture radar (SAR) data which was digitized and recorded on wideband magnetic tape. These tapes were then ground processed into computer compatible tapes (CCT's). The CCT's may then be processed into high resolution radar images by software on the CYBER computer.

  2. Thematic Mapper image quality: Preliminary results

    NASA Technical Reports Server (NTRS)

    Wrigley, R. C.; Card, D. H.; Hlavka, C. A.; Likens, W. C.; Mertz, F. C.; Hall, J. R.

    1983-01-01

    Based on images analyzed so far, the band to band registration accuracy of the thematic mapper is very good. For bands within the same focal plane, the mean misregistrations are well within the specification, 0.2 pixels. For bands between the cooled and uncooled focal planes, there is a consistent mean misregistration of 0.5 pixels along-scan and 0.2-0.3 pixels across-scan. It exceeds the permitted 0.3 pixels for registration of bands between focal planes. If the mean misregistrations were removed by the data processing software, an analysis of the standard deviation of the misregistration indicates all band combinations would meet the registration specifications except for those including the thermal band. Analysis of the periodic noise in one image indicates a noise component in band 1 with a spatial frequency equivalent to 3.2 pixels in the along-scan direction.

  3. Developing Matlab scripts for image analysis and quality assessment

    NASA Astrophysics Data System (ADS)

    Vaiopoulos, A. D.

    2011-11-01

    Image processing is a very helpful tool in many fields of modern sciences that involve digital imaging examination and interpretation. Processed images however, often need to be correlated with the original image, in order to ensure that the resulting image fulfills its purpose. Aside from the visual examination, which is mandatory, image quality indices (such as correlation coefficient, entropy and others) are very useful, when deciding which processed image is the most satisfactory. For this reason, a single program (script) was written in Matlab language, which automatically calculates eight indices by utilizing eight respective functions (independent function scripts). The program was tested in both fused hyperspectral (Hyperion-ALI) and multispectral (ALI, Landsat) imagery and proved to be efficient. Indices were found to be in agreement with visual examination and statistical observations.

  4. TU-A-18C-01: ACR Accreditation Updates in CT, Ultrasound, Mammography and MRI

    SciTech Connect

    Price, R; Berns, E; Hangiandreou, N; McNitt-Gray, M

    2014-06-15

    A goal of an imaging accreditation program is to ensure adequate image quality, verify appropriate staff qualifications, and to assure patient and personnel safety. Currently, more than 35,000 facilities in 10 modalities have been accredited by the American College of Radiology (ACR), making the ACR program one of the most prolific accreditation options in the U.S. In addition, the ACR is one of the accepted accreditations required by some state laws, CMS/MIPPA insurance and others. Familiarity with the ACR accreditation process is therefore essential to clinical diagnostic medical physicists. Maintaining sufficient knowledge of the ACR program must include keeping up-to-date as the various modality requirements are refined to better serve the goals of the program and to accommodate newer technologies and practices. This session consists of presentations from authorities in four ACR accreditation modality programs, including magnetic resonance imaging, mammography, ultrasound, and computed tomography. Each speaker will discuss the general components of the modality program and address any recent changes to the requirements. Learning Objectives: To understand the requirements of the ACR MR accreditation program. The discussion will include accreditation of whole-body general purpose magnets, dedicated extremity systems well as breast MRI accreditation. Anticipated updates to the ACR MRI Quality Control Manual will also be reviewed. To understand the current ACR MAP Accreditation requirement and present the concepts and structure of the forthcoming ACR Digital Mammography QC Manual and Program. To understand the new requirements of the ACR ultrasound accreditation program, and roles the physicist can play in annual equipment surveys and setting up and supervising the routine QC program. To understand the requirements of the ACR CT accreditation program, including updates to the QC manual as well as updates through the FAQ process.

  5. High dynamic range image compression by optimizing tone mapped image quality index.

    PubMed

    Ma, Kede; Yeganeh, Hojatollah; Zeng, Kai; Wang, Zhou

    2015-10-01

    Tone mapping operators (TMOs) aim to compress high dynamic range (HDR) images to low dynamic range (LDR) ones so as to visualize HDR images on standard displays. Most existing TMOs were demonstrated on specific examples without being thoroughly evaluated using well-designed and subject-validated image quality assessment models. A recently proposed tone mapped image quality index (TMQI) made one of the first attempts on objective quality assessment of tone mapped images. Here, we propose a substantially different approach to design TMO. Instead of using any predefined systematic computational structure for tone mapping (such as analytic image transformations and/or explicit contrast/edge enhancement), we directly navigate in the space of all images, searching for the image that optimizes an improved TMQI. In particular, we first improve the two building blocks in TMQI—structural fidelity and statistical naturalness components—leading to a TMQI-II metric. We then propose an iterative algorithm that alternatively improves the structural fidelity and statistical naturalness of the resulting image. Numerical and subjective experiments demonstrate that the proposed algorithm consistently produces better quality tone mapped images even when the initial images of the iteration are created by the most competitive TMOs. Meanwhile, these results also validate the superiority of TMQI-II over TMQI. PMID:26011881

  6. Fast vector quantization algorithm preserving color image quality

    NASA Astrophysics Data System (ADS)

    Charrier, Christophe; Cherifi, Hocine

    1998-04-01

    In the color image compression field, it is well known by researchers that the information is statistically redundant. This redundancy is a handicap in terms of dictionary construction time. A way to counterbalance this time consuming effect is to reduce the redundancy within the original image while keeping the image quality. One can extract a random sample of the initial training set on which one constructs the codebook whose quality is equal to the quality of the codebook generated from the entire training set. We applied this idea in the color vector quantization (VQ) compression scheme context. We propose an algorithm to reduce the complexity of the standard LBG technique. We searched for a measure of relevance of each block from the entire training set. Under the assumption that the measure of relevance is a independent random variable, we applied the Kolmogorov statistical test to define the smallest size of a random sample, and then the sample itself. Finally, from blocks associated to each measure of relevance of the random sample, we compute the standard LBG algorithm to construct the codebook. Psychophysics and statistical measures of image quality allow us to find the best measure of relevance to reduce the training set while preserving the image quality and decreasing the computational cost.

  7. Psychovisual quality metric based on multiscale image texture analysis

    NASA Astrophysics Data System (ADS)

    Eude, Thierry; Mayache, Abdelhalim; Milan, Claude

    1999-05-01

    This paper describes a perceptual measure for still image compression system. Considering the fact that the conventional PSNR cannot sufficiently reflect the result of subjective assessment, other quality measure have been considered to design the variable bit-rate coders. Indeed, there is a growing interest for perceptual quality measure. Some works have been carried out in the field of still picture quality evaluation while trying to introduce some properties of the human visual system. In the recent literature there are roughly three properties that are identified as being useful. The best known, and generally most widely used properly, is the modulation transfer function of the human visual system. The other tow properties can be described as luminance masking and texture masking. A large number of image quality measures of this kind have been developed, with different degrees of success. In previous works, we provided a rigorous evaluation of metrics which take into account artifacts generated by compression method like JPEG. The results show that these metrics are highly correlated with the subjective quality grading but also depend on the complexity of the images under study. Then, we propose a new perceptual metric for still image compression based on multiresolution decomposition. It allows characterize image texture, better takes into account masking effect and don't depend on compression method.

  8. Body image and quality of life in a Spanish population

    PubMed Central

    Lobera, Ignacio Jáuregui; Ríos, Patricia Bolaños

    2011-01-01

    Purpose The aim of the current study was to analyze the psychometric properties, factor structure, and internal consistency of the Spanish version of the Body Image Quality of Life Inventory (BIQLI-SP) as well as its test–retest reliability. Further objectives were to analyze different relationships with key dimensions of psychosocial functioning (ie, self-esteem, presence of psychopathological symptoms, eating and body image-related problems, and perceived stress) and to evaluate differences in body image quality of life due to gender. Patients and methods The sample comprised 417 students without any psychiatric history, recruited from the Pablo de Olavide University and the University of Seville. There were 140 men (33.57%) and 277 women (66.43%), and the mean age was 21.62 years (standard deviation = 5.12). After obtaining informed consent from all participants, the following questionnaires were administered: BIQLI, Eating Disorder Inventory-2 (EDI-2), Perceived Stress Questionnaire (PSQ), Self-Esteem Scale (SES), and Symptom Checklist-90-Revised (SCL-90-R). Results The BIQLI-SP shows adequate psychometric properties, and it may be useful to determine the body image quality of life in different physical conditions. A more positive body image quality of life is associated with better self-esteem, better psychological wellbeing, and fewer eating-related dysfunctional attitudes, this being more evident among women. Conclusion The BIQLI-SP may be useful to determine the body image quality of life in different contexts with regard to dermatology, cosmetic and reconstructive surgery, and endocrinology, among others. In these fields of study, a new trend has emerged to assess body image-related quality of life. PMID:21403794

  9. Slider-adjusted softcopy ruler for calibrated image quality assessment

    NASA Astrophysics Data System (ADS)

    Jin, Elaine W.; Keelan, Brian W.

    2010-01-01

    ISO 20462 part 3 standardized the hardcopy quality ruler and a softcopy quality ruler based on a binary sort approach involving paired comparisons. The new softcopy ruler method described here utilizes a slider bar to match the quality of the ruler to that of the test image, which is found to substantially reduce the time required per assessment (30 to 15.5 s), with only a modest loss of precision (standard deviations of 2.5 to 2.9 just noticeable differences). In combination, these metrics implied a 20% improvement in the standard error of the mean achievable in a fixed amount of judging time. Ruler images calibrated against the standard quality scale of ISO 20462 are generated for 21 scenes, at 31 quality levels each, achieved through variation of sharpness, while other attributes are held near their preferred positions. The images are bundled with documentation and a MATLAB source code for a graphical user interface that administers softcopy ruler experiments, and these materials are donated to the International Imaging Industry Association for distribution. In conjunction with a specified large flat panel display, these materials should enable users to conduct softcopy quality ruler experiments with minimum effort, and should reduce the barriers to performing calibrated psychophysical measurements.

  10. Fast Tomography: A Study Of Image Quality

    SciTech Connect

    Waske, A.; Rahn, H.; Odenbach, S.

    2010-04-06

    We present tomographic slices of fast X-ray computed microtomography, focusing on one hand on the comparison of scans with identical scan time but varying experimental settings, on the other hand on comparing scans with total scan times that are two orders of magnitude apart. We show that streaks in the Fourier-space of the transformed images appear when the number of projections is reduced. These streaks exhibit spacing by the same angle as the one that is covered by the sample during the exposure time of one projection. Further a method for the estimation of information loss in fast scans is presented.

  11. Laboratory validation of a sparse aperture image quality model

    NASA Astrophysics Data System (ADS)

    Salvaggio, Philip S.; Schott, John R.; McKeown, Donald M.

    2015-09-01

    The majority of image quality studies in the field of remote sensing have been performed on systems with conventional aperture functions. These systems have well-understood image quality tradeoffs, characterized by the General Image Quality Equation (GIQE). Advanced, next-generation imaging systems present challenges to both post-processing and image quality prediction. Examples include sparse apertures, synthetic apertures, coded apertures and phase elements. As a result of the non-conventional point spread functions of these systems, post-processing becomes a critical step in the imaging process and artifacts arise that are more complicated than simple edge overshoot. Previous research at the Rochester Institute of Technology's Digital Imaging and Remote Sensing Laboratory has resulted in a modeling methodology for sparse and segmented aperture systems, the validation of which will be the focus of this work. This methodology has predicted some unique post-processing artifacts that arise when a sparse aperture system with wavefront error is used over a large (panchromatic) spectral bandpass. Since these artifacts are unique to sparse aperture systems, they have not yet been observed in any real-world data. In this work, a laboratory setup and initial results for a model validation study will be described. Initial results will focus on the validation of spatial frequency response predictions and verification of post-processing artifacts. The goal of this study is to validate the artifact and spatial frequency response predictions of this model. This will allow model predictions to be used in image quality studies, such as aperture design optimization, and the signal-to-noise vs. post-processing artifact tradeoff resulting from choosing a panchromatic vs. multispectral system.

  12. 21 CFR 892.1715 - Full-field digital mammography system.

    Code of Federal Regulations, 2014 CFR

    2014-04-01

    ... planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports,...

  13. 21 CFR 892.1715 - Full-field digital mammography system.

    Code of Federal Regulations, 2012 CFR

    2012-04-01

    ... planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports,...

  14. 21 CFR 892.1715 - Full-field digital mammography system.

    Code of Federal Regulations, 2013 CFR

    2013-04-01

    ... planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports,...

  15. 21 CFR 892.1715 - Full-field digital mammography system.

    Code of Federal Regulations, 2011 CFR

    2011-04-01

    ... planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports,...

  16. Novel image fusion quality metrics based on sensor models and image statistics

    NASA Astrophysics Data System (ADS)

    Smith, Forrest A.; Chari, Srikant; Halford, Carl E.; Fanning, Jonathan; Reynolds, Joseph P.

    2009-05-01

    This paper presents progress in image fusion modeling. One fusion quality metric based on the Targeting Task performance (TTP) metric and another based on entropy are presented. A human perception test was performed with fused imagery to determine effectiveness of the metrics in predicting image fusion quality. Both fusion metrics first establish which of two source images is ideal in a particular spatial frequency pass band. The fused output of a given algorithm is then measured against this ideal in each pass band. The entropy based fusion quality metric (E-FQM) uses statistical information (entropy) from the images while the Targeting Task Performance fusion quality metric (TTPFQM) utilizes the TTP metric value in each spatial frequency band. This TTP metric value is the measure of available excess contrast determined by the Contrast Threshold Function (CTF) of the source system and the target contrast. The paper also proposes an image fusion algorithm that chooses source image contributions using a quality measure similar to the TTP-FQM. To test the effectiveness of TTP-FQM and E-FQM in predicting human image quality preferences, SWIR and LWIR imagery of tanks were fused using four different algorithms. A paired comparison test was performed with both source and fused imagery as stimuli. Eleven observers were asked to select which image enabled them to better identify the target. Over the ensemble of test images, the experiment showed that both TTP-FQM and E-FQM were capable of identifying the fusion algorithms most and least preferred by human observers. Analysis also showed that the performance of the TTP-FQM and E-FQM in identifying human image preferences are better than existing fusion quality metrics such as the Weighted Fusion Quality Index and Mutual Information.

  17. Perceptual image quality in normalized LOG domain for Adaptive Optics image post-processing

    NASA Astrophysics Data System (ADS)

    Guo, Shiping; Zhang, Rongzhi; Li, Jisheng; Zou, Jianhua; Liu, Changhai; Gao, Weizhe

    2015-08-01

    Adaptive Optics together with subsequent post-processing techniques obviously improve the resolution of turbulencedegraded images in ground-based space objects detection and identification. The most common method for frame selection and stopping iteration in post-processing has always been subjective viewing of the images due to a lack of widely agreed-upon objective quality metric. Full reference metrics are not applicable for assessing the field data, no-reference metrics tend to perform poor sensitivity for Adaptive Optics images. In the present work, based on the Laplacian of Gaussian (LOG) local contrast feature, a nonlinear normalization is applied to transform the input image into a normalized LOG domain; a quantitative index is then extracted in this domain to assess the perceptual image quality. Experiments show this no-reference quality index is highly consistent with the subjective evaluation of input images for different blur degree and different iteration number.

  18. Toward the development of an image quality tool for active millimeter wave imaging systems

    NASA Astrophysics Data System (ADS)

    Barber, Jeffrey; Weatherall, James C.; Greca, Joseph; Smith, Barry T.

    2015-05-01

    Preliminary design considerations for an image quality tool to complement millimeter wave imaging systems are presented. The tool is planned for use in confirming operating parameters; confirmation of continuity for imaging component design changes, and analysis of new components and detection algorithms. Potential embodiments of an image quality tool may contain materials that mimic human skin in order to provide a realistic signal return for testing, which may also help reduce or eliminate the need for mock passengers for developmental testing. Two candidate materials, a dielectric liquid and an iron-loaded epoxy, have been identified and reflection measurements have been performed using laboratory systems in the range 18 - 40 GHz. Results show good agreement with both laboratory and literature data on human skin, particularly in the range of operation of two commercially available millimeter wave imaging systems. Issues related to the practical use of liquids and magnetic materials for image quality tools are discussed.

  19. Diffraction enhanced x-ray imaging

    SciTech Connect

    Thomlinson, W.; Zhong, Z.; Chapman, D.; Johnston, R.E.; Sayers, D.

    1997-09-01

    Diffraction enhanced imaging (DEI) is a new x-ray radiographic imaging modality using synchrotron x-rays which produces images of thick absorbing objects that are almost completely free of scatter. They show dramatically improved contrast over standard imaging applied to the same phantoms. The contrast is based not only on attenuation but also the refraction and diffraction properties of the sample. The diffraction component and the apparent absorption component (absorption plus extinction contrast) can each be determined independently. This imaging method may improve the image quality for medical applications such as mammography.

  20. Single photon counter for digital x-ray mammography tomosynthesis

    SciTech Connect

    Goldan, Amir H.; Karim, Karim S.; Rowlands, John A.

    2006-05-15

    Photon counting is an emerging detection technique that is promising for mammography tomosynthesis imagers. In photon counting systems, the value of each image pixel is equal to the number of photons that interact with the detector. In this research, we introduce the design and implementation of a low noise, photon counting pixel for digital mammography tomosynthesis in 0.18 {mu}m crystalline silicon complementary metal-oxide semiconductor technology. The design comprises of a low noise, charge-integrating amplifier, a low offset voltage comparator, a decision-making unit, a mode selector, and a pseudorandom counter. Theoretical calculations and simulation results of linearity, gain, and noise of the photon counting pixel are presented.

  1. Accuracy of Soft-Copy Digital Mammography versus That of Screen-Film Mammography according to Digital Manufacturer: ACRIN DMIST Retrospective Multireader Study1

    PubMed Central

    Hendrick, R. Edward; Cole, Elodia B.; Pisano, Etta D.; Acharyya, Suddhasatta; Marques, Helga; Cohen, Michael A.; Jong, Roberta A.; Mawdsley, Gordon E.; Kanal, Kalpana M.; D'Orsi, Carl J.; Rebner, Murray; Gatsonis, Constantine

    2008-01-01

    Purpose: To retrospectively compare the accuracy for cancer diagnosis of digital mammography with soft-copy interpretation with that of screen-film mammography for each digital equipment manufacturer, by using results of biopsy and follow-up as the reference standard. Materials and Methods: The primary HIPAA-compliant Digital Mammographic Imaging Screening Trial (DMIST) was approved by the institutional review board of each study site, and informed consent was obtained. The approvals and consent included use of data for future HIPAA-compliant retrospective research. The American College of Radiology Imaging Network DMIST collected screening mammography studies performed by using both digital and screen-film mammography in 49 528 women (mean age, 54.6 years; range, 19–92 years). Digital mammography systems from four manufacturers (Fischer, Fuji, GE, and Hologic) were used. For each digital manufacturer, a cancer-enriched reader set of women screened with both digital and screen-film mammography in DMIST was constructed. Each reader set contained all cancer-containing studies known for each digital manufacturer at the time of reader set selection, together with a subset of negative and benign studies. For each reader set, six or 12 experienced radiologists attended two randomly ordered reading sessions 6 weeks apart. Each radiologist identified suspicious findings and rated suspicion of breast cancer in identified lesions by using a seven-point scale. Results were analyzed according to digital manufacturer by using areas under the receiver operating characteristic curve (AUCs), sensitivity, and specificity for soft-copy digital and screen-film mammography. Results for Hologic digital are not presented owing to the fact that few cancer cases were available. The implemented design provided 80% power to detect average AUC differences of 0.09, 0.08, and 0.06 for Fischer, Fuji, and GE, respectively. Results: No significant difference in AUC, sensitivity, or specificity was found between Fischer, Fuji, and GE soft-copy digital and screen-film mammography. Large reader variations occurred with each modality. Conclusion: No statistically significant differences were found between soft-copy digital and screen-film mammography for Fischer, Fuji, and GE digital mammography equipment. © RSNA, 2008 PMID:18372463

  2. Postmortem validation of breast density using dual-energy mammography

    SciTech Connect

    Molloi, Sabee Ducote, Justin L.; Ding, Huanjun; Feig, Stephen A.

    2014-08-15

    Purpose: Mammographic density has been shown to be an indicator of breast cancer risk and also reduces the sensitivity of screening mammography. Currently, there is no accepted standard for measuring breast density. Dual energy mammography has been proposed as a technique for accurate measurement of breast density. The purpose of this study is to validate its accuracy in postmortem breasts and compare it with other existing techniques. Methods: Forty postmortem breasts were imaged using a dual energy mammography system. Glandular and adipose equivalent phantoms of uniform thickness were used to calibrate a dual energy basis decomposition algorithm. Dual energy decomposition was applied after scatter correction to calculate breast density. Breast density was also estimated using radiologist reader assessment, standard histogram thresholding and a fuzzy C-mean algorithm. Chemical analysis was used as the reference standard to assess the accuracy of different techniques to measure breast composition. Results: Breast density measurements using radiologist reader assessment, standard histogram thresholding, fuzzy C-mean algorithm, and dual energy were in good agreement with the measured fibroglandular volume fraction using chemical analysis. The standard error estimates using radiologist reader assessment, standard histogram thresholding, fuzzy C-mean, and dual energy were 9.9%, 8.6%, 7.2%, and 4.7%, respectively. Conclusions: The results indicate that dual energy mammography can be used to accurately measure breast density. The variability in breast density estimation using dual energy mammography was lower than reader assessment rankings, standard histogram thresholding, and fuzzy C-mean algorithm. Improved quantification of breast density is expected to further enhance its utility as a risk factor for breast cancer.

  3. Postmortem validation of breast density using dual-energy mammography

    PubMed Central

    Molloi, Sabee; Ducote, Justin L.; Ding, Huanjun; Feig, Stephen A.

    2014-01-01

    Purpose: Mammographic density has been shown to be an indicator of breast cancer risk and also reduces the sensitivity of screening mammography. Currently, there is no accepted standard for measuring breast density. Dual energy mammography has been proposed as a technique for accurate measurement of breast density. The purpose of this study is to validate its accuracy in postmortem breasts and compare it with other existing techniques. Methods: Forty postmortem breasts were imaged using a dual energy mammography system. Glandular and adipose equivalent phantoms of uniform thickness were used to calibrate a dual energy basis decomposition algorithm. Dual energy decomposition was applied after scatter correction to calculate breast density. Breast density was also estimated using radiologist reader assessment, standard histogram thresholding and a fuzzy C-mean algorithm. Chemical analysis was used as the reference standard to assess the accuracy of different techniques to measure breast composition. Results: Breast density measurements using radiologist reader assessment, standard histogram thresholding, fuzzy C-mean algorithm, and dual energy were in good agreement with the measured fibroglandular volume fraction using chemical analysis. The standard error estimates using radiologist reader assessment, standard histogram thresholding, fuzzy C-mean, and dual energy were 9.9%, 8.6%, 7.2%, and 4.7%, respectively. Conclusions: The results indicate that dual energy mammography can be used to accurately measure breast density. The variability in breast density estimation using dual energy mammography was lower than reader assessment rankings, standard histogram thresholding, and fuzzy C-mean algorithm. Improved quantification of breast density is expected to further enhance its utility as a risk factor for breast cancer. PMID:25086548

  4. Investigation of perceptual attributes for mobile display image quality

    NASA Astrophysics Data System (ADS)

    Gong, Rui; Xu, Haisong; Wang, Qing; Wang, Zhehong; Li, Haifeng

    2013-08-01

    Large-scale psychophysical experiments are carried out on two types of mobile displays to evaluate the perceived image quality (IQ). Eight perceptual attributes, i.e., naturalness, colorfulness, brightness, contrast, sharpness, clearness, preference, and overall IQ, are visually assessed via categorical judgment method for various application types of test images, which were manipulated by different methods. Their correlations are deeply discussed, and further factor analysis revealed the two essential components to describe the overall IQ, i.e., the component of image detail aspect and the component of color information aspect. Clearness and naturalness are regarded as two principal factors for natural scene images, whereas clearness and colorfulness were selected as key attributes affecting the overall IQ for other application types of images. Accordingly, based on these selected attributes, two kinds of empirical models are built to predict the overall IQ of mobile displays for different application types of images.

  5. The influence of noise on image quality in phase-diverse coherent diffraction imaging

    NASA Astrophysics Data System (ADS)

    Wittler, H. P. A.; van Riessen, G. A.; Jones, M. W. M.

    2016-02-01

    Phase-diverse coherent diffraction imaging provides a route to high sensitivity and resolution with low radiation dose. To take full advantage of this, the characteristics and tolerable limits of measurement noise for high quality images must be understood. In this work we show the artefacts that manifest in images recovered from simulated data with noise of various characteristics in the illumination and diffraction pattern. We explore the limits at which images of acceptable quality can be obtained and suggest qualitative guidelines that would allow for faster data acquisition and minimize radiation dose.

  6. Breast cancer detection and classification in digital mammography based on Non-Subsampled Contourlet Transform (NSCT) and Super Resolution.

    PubMed

    Pak, Fatemeh; Kanan, Hamidreza Rashidy; Alikhassi, Afsaneh

    2015-11-01

    Breast cancer is one of the most perilous diseases among women. Breast screening is a method of detecting breast cancer at a very early stage which can reduce the mortality rate. Mammography is a standard method for the early diagnosis of breast cancer. In this paper, a new algorithm is proposed for breast cancer detection and classification in digital mammography based on Non-Subsampled Contourlet Transform (NSCT) and Super Resolution (SR). The presented algorithm includes three main parts including pre-processing, feature extraction and classification. In the pre-processing stage, after determining the region of interest (ROI) by an automatic technique, the quality of image is improved using NSCT and SR algorithm. In the feature extraction part, several features of the image components are extracted and skewness of each feature is calculated. Finally, AdaBoost algorithm is used to classify and determine the probability of benign and malign disease. The obtained results on Mammographic Image Analysis Society (MIAS) database indicate the significant performance and superiority of the proposed method in comparison with the state of the art approaches. According to the obtained results, the proposed technique achieves 91.43% and 6.42% as a mean accuracy and FPR, respectively. PMID:26206406

  7. AEC for scanning digital mammography based on variation of scan velocity

    SciTech Connect

    Aaslund, Magnus; Cederstroem, Bjoern; Lundqvist, Mats; Danielsson, Mats

    2005-11-15

    A theoretical evaluation of nonuniform x-ray field distributions in mammography was conducted. An automatic exposure control (AEC) is proposed for a scanning full field digital mammography system. It uses information from the leading part of the detector to vary the scan velocity dynamically, thus creating a nonuniform x-ray field in the scan direction. Nonuniform radiation fields were also created by numerically optimizing the scan velocity profile to each breast's transmission distribution, with constraints on velocity and acceleration. The goal of the proposed AEC is to produce constant pixel signal-to-noise ratio throughout the image. The target pixel SNR for each image could be set based on the breast thickness, breast composition, and the beam quality as to achieve the same contrast-to-noise ratio between images for structures of interest. The results are quantified in terms of reduction in entrance surface air kerma (ESAK) and scan time relative to a uniform x-ray field. The theoretical evaluation was performed on a set of 266 mammograms. The performance of the different methods to create nonuniform fields decreased with increased detector width, from 18% to 11% in terms of ESAK reduction and from 30% to 25% in terms of scan time reduction for the proposed AEC and detector widths from 10 to 60 mm. Some correlation was found between compressed breast thickness and the projected breast area onto the image field. This translated into an increase of the ESAK and decrease of the scan time reduction with breast thickness. Ideally a nonuniform field in two dimensions could reduce the entrance dose by 39% on average, whereas a field nonuniform in only the scanning dimension ideally yields a 20% reduction. A benefit with the proposed AEC is that the risk of underexposing the densest region of the breast can be virtually eliminated.

  8. Real-time computer treatment of THz passive device images with the high image quality

    NASA Astrophysics Data System (ADS)

    Trofimov, Vyacheslav A.; Trofimov, Vladislav V.

    2012-06-01

    We demonstrate real-time computer code improving significantly the quality of images captured by the passive THz imaging system. The code is not only designed for a THz passive device: it can be applied to any kind of such devices and active THz imaging systems as well. We applied our code for computer processing of images captured by four passive THz imaging devices manufactured by different companies. It should be stressed that computer processing of images produced by different companies requires using the different spatial filters usually. The performance of current version of the computer code is greater than one image per second for a THz image having more than 5000 pixels and 24 bit number representation. Processing of THz single image produces about 20 images simultaneously corresponding to various spatial filters. The computer code allows increasing the number of pixels for processed images without noticeable reduction of image quality. The performance of the computer code can be increased many times using parallel algorithms for processing the image. We develop original spatial filters which allow one to see objects with sizes less than 2 cm. The imagery is produced by passive THz imaging devices which captured the images of objects hidden under opaque clothes. For images with high noise we develop an approach which results in suppression of the noise after using the computer processing and we obtain the good quality image. With the aim of illustrating the efficiency of the developed approach we demonstrate the detection of the liquid explosive, ordinary explosive, knife, pistol, metal plate, CD, ceramics, chocolate and other objects hidden under opaque clothes. The results demonstrate the high efficiency of our approach for the detection of hidden objects and they are a very promising solution for the security problem.

  9. Physical characteristics of five clinical systems for digital mammography

    SciTech Connect

    Lazzari, B.; Belli, G.; Gori, C.; Rosselli Del Turco, M.

    2007-07-15

    The purpose of this study was to evaluate and compare the physical characteristics of five clinical systems for digital mammography (GE Senographe 2000D, Lorad Selenia M-IV, Fischer Senoscan, Agfa DM 1000, and IMS Giotto) currently in clinical use. The basic performances of the mammography systems tested were assessed on the basis of response curve, modulation transfer function (MTF), noise power spectrum, noise equivalent quanta (NEQ), and detective quantum efficiency (DQE) in an experimental setting closely resembling the clinical one. As expected, all the full field digital mammography systems show a linear response curve over a dynamic range from 3.5 to 500 {mu}Gy (0.998image quality. The detailed results of the physical characterization of the digital systems reported in this work allow the quantitative comparison of different technologies as well as the definition of reference values for subsequent quality control tests. The method developed in this work is suitable to be reproduced in any medical physics department for the previously described goals.

  10. Physical characteristics of five clinical systems for digital mammography.

    PubMed

    Lazzari, B; Belli, G; Gori, C; Rosselli Del Turco, M

    2007-07-01

    The purpose of this study was to evaluate and compare the physical characteristics of five clinical systems for digital mammography (GE Senographe 2000D, Lorad Selenia M-IV, Fischer Senoscan, Agfa DM 1000, and IMS Giotto) currently in clinical use. The basic performances of the mammography systems tested were assessed on the basis of response curve, modulation transfer function (MTF), noise power spectrum, noise equivalent quanta (NEQ), and detective quantum efficiency (DQE) in an experimental setting closely resembling the clinical one. As expected, all the full field digital mammography systems show a linear response curve over a dynamic range from 3.5 to 500 microGy (0.998image quality. The detailed results of the physical characterization of the digital systems reported in this work allow the quantitative comparison of different technologies as well as the definition of reference values for subsequent quality control tests. The method developed in this work is suitable to be reproduced in any medical physics department for the previously described goals. PMID:17821981

  11. X-ray source for mammography

    DOEpatents

    Logan, C.M.

    1994-12-20

    An x-ray source is described utilizing anode material which shifts the output spectrum to higher energy and thereby obtains higher penetrating ability for screening mammography application, than the currently utilized anode material. The currently used anode material (molybdenum) produces an energy x-ray spectrum of 17.5/19.6 keV, which using the anode material of this invention (e.g. silver, rhodium, and tungsten) the x-ray spectrum would be in the 20-35 keV region. Thus, the anode material of this invention provides for imaging of breasts with higher than average x-ray opacity without increase of the radiation dose, and thus reduces the risk of induced breast cancer due to the radiation dose administered for mammograms. 6 figures.

  12. X-ray source for mammography

    DOEpatents

    Logan, Clinton M.

    1994-01-01

    An x-ray source utilizing anode material which shifts the output spectrum to higher energy and thereby obtains higher penetrating ability for screening mammography application, than the currently utilized anode material. The currently used anode material (molybdenum) produces an energy x-ray spectrum of 17.5/19.6 keV, which using the anode material of this invention (e.g. silver, rhodium, and tungsten) the x-ray spectrum would be in the 20-35 keV region. Thus, the anode material of this invention provides for imaging of breasts with higher than average x-ray opacity without increase of the radiation dose, and thus reduces the risk of induced breast cancer due to the radiation dose administered for mammograms.

  13. Instrumentation optimization for positron emission mammography

    SciTech Connect

    Moses, William W.; Qi, Jinyi

    2003-06-05

    The past several years have seen designs for PET cameras optimized to image the breast, commonly known as Positron Emission Mammography or PEM cameras. The guiding principal behind PEM instrumentation is that a camera whose field of view is restricted to a single breast has higher performance and lower cost than a conventional PET camera. The most common geometry is a pair of parallel planes of detector modules, although geometries that encircle the breast have also been proposed. The ability of the detector modules to measure the depth of interaction (DOI) is also a relevant feature. This paper finds that while both the additional solid angle coverage afforded by encircling the breast and the decreased blurring afforded by the DOI measurement improve performance, the ability to measure DOI is more important than the ability to encircle the breast.

  14. Determination of pasture quality using airborne hyperspectral imaging

    NASA Astrophysics Data System (ADS)

    Pullanagari, R. R.; Kereszturi, G.; Yule, Ian J.; Irwin, M. E.

    2015-10-01

    Pasture quality is a critical determinant which influences animal performance (live weight gain, milk and meat production) and animal health. Assessment of pasture quality is therefore required to assist farmers with grazing planning and management, benchmarking between seasons and years. Traditionally, pasture quality is determined by field sampling which is laborious, expensive and time consuming, and the information is not available in real-time. Hyperspectral remote sensing has potential to accurately quantify biochemical composition of pasture over wide areas in great spatial detail. In this study an airborne imaging spectrometer (AisaFENIX, Specim) was used with a spectral range of 380-2500 nm with 448 spectral bands. A case study of a 600 ha hill country farm in New Zealand is used to illustrate the use of the system. Radiometric and atmospheric corrections, along with automatized georectification of the imagery using Digital Elevation Model (DEM), were applied to the raw images to convert into geocoded reflectance images. Then a multivariate statistical method, partial least squares (PLS), was applied to estimate pasture quality such as crude protein (CP) and metabolisable energy (ME) from canopy reflectance. The results from this study revealed that estimates of CP and ME had a R2 of 0.77 and 0.79, and RMSECV of 2.97 and 0.81 respectively. By utilizing these regression models, spatial maps were created over the imaged area. These pasture quality maps can be used for adopting precision agriculture practices which improves farm profitability and environmental sustainability.

  15. Perceived assessment metrics for visible and infrared color fused image quality without reference image

    NASA Astrophysics Data System (ADS)

    Yu, Xuelian; Chen, Qian; Gu, Guohua; Ren, Jianle; Sui, Xiubao

    2015-02-01

    Designing objective quality assessment of color-fused image is a very demanding and challenging task. We propose four no-reference metrics based on human visual system characteristics for objectively evaluating the quality of false color fusion image. The perceived edge metric (PEM) is defined based on visual perception model and color image gradient similarity between the fused image and the source images. The perceptual contrast metric (PCM) is established associating multi-scale contrast and varying contrast sensitivity filter (CSF) with color components. The linear combination of the standard deviation and mean value over the fused image construct the image colorfulness metric (ICM). The color comfort metric (CCM) is designed by the average saturation and the ratio of pixels with high and low saturation. The qualitative and quantitative experimental results demonstrate that the proposed metrics have a good agreement with subjective perception.

  16. Diagnosing breast cancer using independent diffuse optical tomography and x-ray mammography scans

    NASA Astrophysics Data System (ADS)

    Fradkin, Maxim; Hofmann, Matthias C.; Rouet, Jean-Michel; Moore, Richard H.; Kopans, Daniel B.; Tipton, Keith; Suryanarayanan, Sankar; Boas, David A.; Fang, Qianqian

    2013-03-01

    We have previously demonstrated the utilization of spatially co-registered diffuse optical tomography (DOT) and digital breast tomosynthesis (DBT) for joint breast cancer diagnosis. However, clinical implementation of such a multi-modality approach may require development of integrated DOT/DBT imaging scanners, which can be costly and time-consuming. Exploring effective image registration methods that combine the diagnostic information from a standalone DOT measurement and a separate mammogram can be a cost-effective solution, which may eventually enable adding functional optical assessment to all previously installed digital mammography systems. In this study, we investigate a contour-based image registration method to convert independent optical and x-ray scans into co-registered datasets that can benefit from a joint image analysis. The breast surface used in 3D optical DOT reconstruction is registered with the breast contour line extracted from an x-ray mammogram acquired separately. This allows us to map the 2D mammogram to the optical measurement space and build structural constraints for optical image reconstruction. A non-linear reconstruction utilizing structure-priors is then performed to produce hemoglobin maps with improved resolution. To validate this approach, we used a set of tumor patient measurements with simultaneous DOT/DBT and separate 2D mammographic scans. The images recovered from the registration procedure derived from DOT and 2D mammogram present similar image quality compared to those recovered from co-registered DOT/DBT measurements.

  17. Objective quality evaluation of visible and infrared color fusion image

    NASA Astrophysics Data System (ADS)

    Yuan, Yihui; Zhang, Junju; Chang, Benkang; Han, Yiyong

    2011-03-01

    An evaluation for objectively assessing the quality of visible and infrared color fusion image is proposed. On the basis of the consideration that human perception is most sensitive to color, sharpness, and contrast when assessing the quality of color image, we propose four objective metrics: image sharpness metric (ISM), image contrast metric (ICM), color colorfulness metric (CCM), and color naturalness metric (CNM). The ISM is evaluated by image gradient information. The ICM is defined based on both gray and color histogram characteristics. A color chroma metric, as well as a color variety metric based on a color difference gradient, is proposed, respectively, to define the CCM. The CNM is defined by measuring the color distribution's similarity between the fusion image and nature image, which are of the same scene. All the color attributions are computed in the CIELAB color space. Experimental results show that the proposed objective metrics are meaningful and effective on color fusion image evaluation because they correspond well to subjective evaluation.

  18. Radiation dose and image quality for paediatric interventional cardiology

    NASA Astrophysics Data System (ADS)

    Vano, E.; Ubeda, C.; Leyton, F.; Miranda, P.

    2008-08-01

    Radiation dose and image quality for paediatric protocols in a biplane x-ray system used for interventional cardiology have been evaluated. Entrance surface air kerma (ESAK) and image quality using a test object and polymethyl methacrylate (PMMA) phantoms have been measured for the typical paediatric patient thicknesses (4-20 cm of PMMA). Images from fluoroscopy (low, medium and high) and cine modes have been archived in digital imaging and communications in medicine (DICOM) format. Signal-to-noise ratio (SNR), figure of merit (FOM), contrast (CO), contrast-to-noise ratio (CNR) and high contrast spatial resolution (HCSR) have been computed from the images. Data on dose transferred to the DICOM header have been used to test the values of the dosimetric display at the interventional reference point. ESAK for fluoroscopy modes ranges from 0.15 to 36.60 µGy/frame when moving from 4 to 20 cm PMMA. For cine, these values range from 2.80 to 161.10 µGy/frame. SNR, FOM, CO, CNR and HCSR are improved for high fluoroscopy and cine modes and maintained roughly constant for the different thicknesses. Cumulative dose at the interventional reference point resulted 25-45% higher than the skin dose for the vertical C-arm (depending of the phantom thickness). ESAK and numerical image quality parameters allow the verification of the proper setting of the x-ray system. Knowing the increases in dose per frame when increasing phantom thicknesses together with the image quality parameters will help cardiologists in the good management of patient dose and allow them to select the best imaging acquisition mode during clinical procedures.

  19. Factors that affect print quality in thermal dye transfer imaging

    NASA Astrophysics Data System (ADS)

    Harrison, Daniel J. P.; McInerney, Elizabeth

    1995-04-01

    Thermal dye transfer (TDT) imaging has established itself as the state- of-the-art process for high quality, continuous tone, nonimpact printing. Imaging quality from this process rivals conventional silver halide photography and exceeds other nonimpact printing technologies. Because this output appears to be virtually indistinguishable from photographic prints, there has been an expectation that all the quality attributes of silver halide photography are embodied in a TDT print. However, there are many significant differences that affect output quality between these two technologies. These differences are primarily in color gamut, print artifacts, Dmin, grain/sharpness, and image stability. The range of colors reproducible by a color, hard copy device, known as its color gamut, is dictated primarily by the image- forming dyes used by the device. The size and shape of a device's gamut is controlled by the spectral density distributions of these image forming dyes, the Dmin of the receiver base, the Dmax of each dye, the amount of light scatter, and the spectral distribution of the viewing illuminant. The spectral density distributions of dyes also have an impact on illuminant sensitivity, which is a predictor of how much the color balance of a print will change with a change in illuminant. By determining and then using characteristic curves for various image- forming dyes, we have been able to calculate and compare the color gamuts and illuminant sensitivity of TDT imaging with other technologies (color monitor and silver halide photography, for example). The differences we have found can have a significant impact on output quality, depending upon the application. Compared to conventional photography, thermal dye transfer prints have traditionally had inferior light stability and resistance to damage from fingerprints. In addition, thermal dye transfer prints have been aggressively attacked by plasticized polyvinyl chloride sheets and folders commonly found in office and home environments. We will describe a major advance in thermal dye transfer imaging technology that greatly improves the image stability position of thermal dye transfer images. This advance is derived from the addition of a thin protective layer onto the final print. To add to customer convenience, the protective layer is integrated into the dye donor ribbon as a 4th patch. The protective layer is laminated to the final print using the thermal print head. TDT print artifacts may also influence the quality of TDT output. These defects can include print head streaks, dust and dirt spots, printer banding, and donor ribbon wrinkling. The origin of these defects will be described.

  20. Compressed image quality metric based on perceptually weighted distortion.

    PubMed

    Hu, Sudeng; Jin, Lina; Wang, Hanli; Zhang, Yun; Kwong, Sam; Kuo, C-C Jay

    2015-12-01

    Objective quality assessment for compressed images is critical to various image compression systems that are essential in image delivery and storage. Although the mean squared error (MSE) is computationally simple, it may not be accurate to reflect the perceptual quality of compressed images, which is also affected dramatically by the characteristics of human visual system (HVS), such as masking effect. In this paper, an image quality metric (IQM) is proposed based on perceptually weighted distortion in terms of the MSE. To capture the characteristics of HVS, a randomness map is proposed to measure the masking effect and a preprocessing scheme is proposed to simulate the processing that occurs in the initial part of HVS. Since the masking effect highly depends on the structural randomness, the prediction error from neighborhood with a statistical model is used to measure the significance of masking. Meanwhile, the imperceptible signal with high frequency could be removed by preprocessing with low-pass filters. The relation is investigated between the distortions before and after masking effect, and a masking modulation model is proposed to simulate the masking effect after preprocessing. The performance of the proposed IQM is validated on six image databases with various compression distortions. The experimental results show that the proposed algorithm outperforms other benchmark IQMs. PMID:26415170

  1. Dose and diagnostic image quality in digital tomosynthesis imaging of facial bones in pediatrics

    NASA Astrophysics Data System (ADS)

    King, J. M.; Hickling, S.; Elbakri, I. A.; Reed, M.; Wrogemann, J.

    2011-03-01

    The purpose of this study was to evaluate the use of digital tomosynthesis (DT) for pediatric facial bone imaging. We compared the eye lens dose and diagnostic image quality of DT facial bone exams relative to digital radiography (DR) and computed tomography (CT), and investigated whether we could modify our current DT imaging protocol to reduce patient dose while maintaining sufficient diagnostic image quality. We measured the dose to the eye lens for all three modalities using high-sensitivity thermoluminescent dosimeters (TLDs) and an anthropomorphic skull phantom. To assess the diagnostic image quality of DT compared to the corresponding DR and CT images, we performed an observer study where the visibility of anatomical structures in the DT phantom images were rated on a four-point scale. We then acquired DT images at lower doses and had radiologists indicate whether the visibility of each structure was adequate for diagnostic purposes. For typical facial bone exams, we measured eye lens doses of 0.1-0.4 mGy for DR, 0.3-3.7 mGy for DT, and 26 mGy for CT. In general, facial bone structures were visualized better with DT then DR, and the majority of structures were visualized well enough to avoid the need for CT. DT imaging provides high quality diagnostic images of the facial bones while delivering significantly lower doses to the lens of the eye compared to CT. In addition, we found that by adjusting the imaging parameters, the DT effective dose can be reduced by up to 50% while maintaining sufficient image quality.

  2. Perceived Image Quality Improvements from the Application of Image Deconvolution to Retinal Images from an Adaptive Optics Fundus Imager

    NASA Astrophysics Data System (ADS)

    Soliz, P.; Nemeth, S. C.; Erry, G. R. G.; Otten, L. J.; Yang, S. Y.

    Aim: The objective of this project was to apply an image restoration methodology based on wavefront measurements obtained with a Shack-Hartmann sensor and evaluating the restored image quality based on medical criteria.Methods: Implementing an adaptive optics (AO) technique, a fundus imager was used to achieve low-order correction to images of the retina. The high-order correction was provided by deconvolution. A Shack-Hartmann wavefront sensor measures aberrations. The wavefront measurement is the basis for activating a deformable mirror. Image restoration to remove remaining aberrations is achieved by direct deconvolution using the point spread function (PSF) or a blind deconvolution. The PSF is estimated using measured wavefront aberrations. Direct application of classical deconvolution methods such as inverse filtering, Wiener filtering or iterative blind deconvolution (IBD) to the AO retinal images obtained from the adaptive optical imaging system is not satisfactory because of the very large image size, dificulty in modeling the system noise, and inaccuracy in PSF estimation. Our approach combines direct and blind deconvolution to exploit available system information, avoid non-convergence, and time-consuming iterative processes. Results: The deconvolution was applied to human subject data and resulting restored images compared by a trained ophthalmic researcher. Qualitative analysis showed significant improvements. Neovascularization can be visualized with the adaptive optics device that cannot be resolved with the standard fundus camera. The individual nerve fiber bundles are easily resolved as are melanin structures in the choroid. Conclusion: This project demonstrated that computer-enhanced, adaptive optic images have greater detail of anatomical and pathological structures.

  3. Estimation of the RBE of mammography-quality beams using a combination of a Monte Carlo code with a B-DNA geometrical model

    NASA Astrophysics Data System (ADS)

    Bernal, M. A.; deAlmeida, C. E.; David, M.; Pires, E.

    2011-12-01

    The PENELOPE code is used to determine direct strand break yields corresponding to photons from a 60Co source and 28 and 30 kV x-ray beams impacting on a B-DNA geometrical model, which accounts for five organizational levels of the human genetic material. Direct single, double and total strand break probabilities are determined in a liquid water homogeneous medium with 1.06 g cm-3 density. The spectra produced by the x-ray beams at various depths in the phantom have been used to study the dependence of the damage yield on the depth. The relative biological effectiveness (RBE) is also estimated using the 60Co radiation qualities as the reference. According to this work, the damage probabilities and thus the RBE are, within the uncertainties, similar for both x-ray energies and are independent of the depth into the phantom. Furthermore, the total strand break yield is invariant with respect to the energy of the incident photons. The RBE for low-energy x-ray beams determined here (1.3 ± 0.1) is lower than that reported by Kellerer, taking into account that he used a 200 kV radiation as the reference quality. However, our RBE values are consistent with those determined by Kühne et al (2005 Radiat. Res. 164 669-76), which used the same biological endpoint and reference quality as our study. Also, our RBE values are similar to those determined by Verhaegen and Reniers (2004 Radiat. Res. 162 592-9).

  4. Flattening filter removal for improved image quality of megavoltage fluoroscopy

    SciTech Connect

    Christensen, James D.; Kirichenko, Alexander; Gayou, Olivier

    2013-08-15

    Purpose: Removal of the linear accelerator (linac) flattening filter enables a high rate of dose deposition with reduced treatment time. When used for megavoltage imaging, an unflat beam has reduced primary beam scatter resulting in sharper images. In fluoroscopic imaging mode, the unflat beam has higher photon count per image frame yielding higher contrast-to-noise ratio. The authors’ goal was to quantify the effects of an unflat beam on the image quality of megavoltage portal and fluoroscopic images.Methods: 6 MV projection images were acquired in fluoroscopic and portal modes using an electronic flat-panel imager. The effects of the flattening filter on the relative modulation transfer function (MTF) and contrast-to-noise ratio were quantified using the QC3 phantom. The impact of FF removal on the contrast-to-noise ratio of gold fiducial markers also was studied under various scatter conditions.Results: The unflat beam had improved contrast resolution, up to 40% increase in MTF contrast at the highest frequency measured (0.75 line pairs/mm). The contrast-to-noise ratio was increased as expected from the increased photon flux. The visualization of fiducial markers was markedly better using the unflat beam under all scatter conditions, enabling visualization of thin gold fiducial markers, the thinnest of which was not visible using the unflat beam.Conclusions: The removal of the flattening filter from a clinical linac leads to quantifiable improvements in the image quality of megavoltage projection images. These gains enable observers to more easily visualize thin fiducial markers and track their motion on fluoroscopic images.

  5. Body image quality of life in eating disorders

    PubMed Central

    Jáuregui Lobera, Ignacio; Bolaños Ríos, Patricia

    2011-01-01

    Purpose: The objective was to examine how body image affects quality of life in an eating-disorder (ED) clinical sample, a non-ED clinical sample, and a nonclinical sample. We hypothesized that ED patients would show the worst body image quality of life. We also hypothesized that body image quality of life would have a stronger negative association with specific ED-related variables than with other psychological and psychopathological variables, mainly among ED patients. On the basis of previous studies, the influence of gender on the results was explored, too. Patients and methods: The final sample comprised 70 ED patients (mean age 22.65 ± 7.76 years; 59 women and 11 men); 106 were patients with other psychiatric disorders (mean age 28.20 ± 6.52; 67 women and 39 men), and 135 were university students (mean age 21.57 ± 2.58; 81 women and 54 men), with no psychiatric history. After having obtained informed consent, the following questionnaires were administered: Body Image Quality of Life Inventory-Spanish version (BIQLI-SP), Eating Disorders Inventory-2 (EDI-2), Perceived Stress Questionnaire (PSQ), Self-Esteem Scale (SES), and Symptom Checklist-90-Revised (SCL-90-R). Results: The ED patients’ ratings on the BIQLI-SP were the lowest and negatively scored (BIQLI-SP means: +20.18, +5.14, and ?6.18, in the student group, the non-ED patient group, and the ED group, respectively). The effect of body image on quality of life was more negative in the ED group in all items of the BIQLI-SP. Body image quality of life was negatively associated with specific ED-related variables, more than with other psychological and psychopathological variables, but not especially among ED patients. Conclusion: Body image quality of life was affected not only by specific pathologies related to body image disturbances, but also by other psychopathological syndromes. Nevertheless, the greatest effect was related to ED, and seemed to be more negative among men. This finding is the opposite of that found in other groups studied previously. PMID:21448468

  6. Image quality testing of assembled IR camera modules

    NASA Astrophysics Data System (ADS)

    Winters, Daniel; Erichsen, Patrik

    2013-10-01

    Infrared (IR) camera modules for the LWIR (8-12_m) that combine IR imaging optics with microbolometer focal plane array (FPA) sensors with readout electronics are becoming more and more a mass market product. At the same time, steady improvements in sensor resolution in the higher priced markets raise the requirement for imaging performance of objectives and the proper alignment between objective and FPA. This puts pressure on camera manufacturers and system integrators to assess the image quality of finished camera modules in a cost-efficient and automated way for quality control or during end-of-line testing. In this paper we present recent development work done in the field of image quality testing of IR camera modules. This technology provides a wealth of additional information in contrast to the more traditional test methods like minimum resolvable temperature difference (MRTD) which give only a subjective overall test result. Parameters that can be measured are image quality via the modulation transfer function (MTF) for broadband or with various bandpass filters on- and off-axis and optical parameters like e.g. effective focal length (EFL) and distortion. If the camera module allows for refocusing the optics, additional parameters like best focus plane, image plane tilt, auto-focus quality, chief ray angle etc. can be characterized. Additionally, the homogeneity and response of the sensor with the optics can be characterized in order to calculate the appropriate tables for non-uniformity correction (NUC). The technology can also be used to control active alignment methods during mechanical assembly of optics to high resolution sensors. Other important points that are discussed are the flexibility of the technology to test IR modules with different form factors, electrical interfaces and last but not least the suitability for fully automated measurements in mass production.

  7. High-speed terahertz imaging toward food quality inspection.

    PubMed

    Ok, Gyeongsik; Park, Kisang; Kim, Hyun Jung; Chun, Hyang Sook; Choi, Sung-Wook

    2014-03-01

    In contrast to conventional x-ray food inspection systems that have difficulty in detecting low-density materials, a terahertz imaging system can even identify insects and plastics embedded in a food matrix. A reflection-mode continuous-wave terahertz imaging system was therefore developed for application to food quality inspection, which requires fast, compact, and low-cost detection. High-speed operation of the terahertz imaging system was achieved through the use of a beam-steering tool. A reasonable compromise between the spatial resolution and the scan length of an aspheric f-theta scanning lens could be achieved by optimizing the lens parameters. PMID:24663370

  8. Definition on SAR image quality measurements for UWB SAR

    NASA Astrophysics Data System (ADS)

    Vu, Viet T.; Sjögren, Thomas K.; Pettersson, Mats I.; Gustavsson, Anders

    2008-10-01

    Analyses in this study show that measurements under currently used definitions on SAR image quality measurement may be unsuitable for UWB SAR. The main objective of this paper is therefore to propose a definition based on the shape of a single point target in a SAR image which is more suitable for UWB SAR. We use both real and simulated data based on the airborne UWB low frequency SAR CARABAS-II in experiments. The time-domain algorithm Global Backprojection (GBP) is selected for the image formation in this study.

  9. Pyramid wavefront sensor for image quality evaluation of optical system

    NASA Astrophysics Data System (ADS)

    Chen, Zhendong

    2015-08-01

    When the pyramid wavefront sensor is used to evaluate the imaging quality, placed at the focal plane of the aberrated optical system e.g., a telescope, it splits the light into four beams. Four images of the pupil are created on the detector and the detection signals of the pyramid wavefront sensor are calculated with these four intensity patterns, providing information on the derivatives of the aberrated wavefront. Based on the theory of the pyramid wavefront sensor, we are going to develop simulation software and a wavefront detector which can be used to test the imaging quality of the telescope. In our system, the subpupil image intensity through the pyramid sensor is calculated to obtain the aberration of wavefront where the piston, tilt, defocus, spherical, coma, astigmatism and other high level aberrations are separately represented by Zernike polynomials. The imaging quality of the optical system is then evaluated by the subsequent wavefront reconstruction. The performance of our system is to be checked by comparing with the measurements carried out using Puntino wavefront instrument (the method of SH wavefront sensor). Within this framework, the measurement precision of pyramid sensor will be discussed as well through detailed experiments. In general, this project would be very helpful both in our understanding of the principle of the wavefront reconstruction and its future technical applications. So far, we have produced the pyramid and established the laboratory setup of the image quality detecting system based on this wavefront sensor. Preliminary results are obtained, in that we have obtained the intensity images of the four pupils. Additional work is needed to analyze the characteristics of the pyramid wavefront sensor.

  10. Digital Mammography Example Dataset Documenation

    Cancer.gov

    This dataset includes 20,000 digital and 20,000 film screening mammograms performed between January 2005 and December 2008 from women included in the Breast Cancer Surveillance Consortium. Some women contribute more than one examination to the dataset. These data are recommended for use in teaching data analysis or epidemiological concepts. Because they represent only a small sample of mammography data available from BCSC they should not be used to conduct primary research.

  11. Color image encryption using a high-quality elemental image array

    NASA Astrophysics Data System (ADS)

    Li, Xiao-Wei; Kim, Seok-Tae; Lee, In-Kwon

    2014-12-01

    In this paper, we present a color image encoding algorithm by combined use of the high-quality elemental image array (EIA) and the pseudo-random mask. To overcome low resolution drawbacks in widely used optical pickup system, in our scheme, the pseudo-inverse filter is introduced to improve this problem. In the cryptosystem, the proposed scheme provides high security because of the high key space of cellular automata. Meanwhile, the hologram-like attribute of the EIA can significantly improve the robustness of the encrypted image against some common image processing attacks. Experiments and analysis have both demonstrated the feasibility and efficiency of the image encryption algorithm.

  12. Simultaneous analysis and quality assurance for diffusion tensor imaging.

    PubMed

    Lauzon, Carolyn B; Asman, Andrew J; Esparza, Michael L; Burns, Scott S; Fan, Qiuyun; Gao, Yurui; Anderson, Adam W; Davis, Nicole; Cutting, Laurie E; Landman, Bennett A

    2013-01-01

    Diffusion tensor imaging (DTI) enables non-invasive, cyto-architectural mapping of in vivo tissue microarchitecture through voxel-wise mathematical modeling of multiple magnetic resonance imaging (MRI) acquisitions, each differently sensitized to water diffusion. DTI computations are fundamentally estimation processes and are sensitive to noise and artifacts. Despite widespread adoption in the neuroimaging community, maintaining consistent DTI data quality remains challenging given the propensity for patient motion, artifacts associated with fast imaging techniques, and the possibility of hardware changes/failures. Furthermore, the quantity of data acquired per voxel, the non-linear estimation process, and numerous potential use cases complicate traditional visual data inspection approaches. Currently, quality inspection of DTI data has relied on visual inspection and individual processing in DTI analysis software programs (e.g. DTIPrep, DTI-studio). However, recent advances in applied statistical methods have yielded several different metrics to assess noise level, artifact propensity, quality of tensor fit, variance of estimated measures, and bias in estimated measures. To date, these metrics have been largely studied in isolation. Herein, we select complementary metrics for integration into an automatic DTI analysis and quality assurance pipeline. The pipeline completes in 24 hours, stores statistical outputs, and produces a graphical summary quality analysis (QA) report. We assess the utility of this streamlined approach for empirical quality assessment on 608 DTI datasets from pediatric neuroimaging studies. The efficiency and accuracy of quality analysis using the proposed pipeline is compared with quality analysis based on visual inspection. The unified pipeline is found to save a statistically significant amount of time (over 70%) while improving the consistency of QA between a DTI expert and a pool of research associates. Projection of QA metrics to a low dimensional manifold reveal qualitative, but clear, QA-study associations and suggest that automated outlier/anomaly detection would be feasible. PMID:23637895

  13. Simultaneous Analysis and Quality Assurance for Diffusion Tensor Imaging

    PubMed Central

    Lauzon, Carolyn B.; Asman, Andrew J.; Esparza, Michael L.; Burns, Scott S.; Fan, Qiuyun; Gao, Yurui; Anderson, Adam W.; Davis, Nicole; Cutting, Laurie E.; Landman, Bennett A.

    2013-01-01

    Diffusion tensor imaging (DTI) enables non-invasive, cyto-architectural mapping of in vivo tissue microarchitecture through voxel-wise mathematical modeling of multiple magnetic resonance imaging (MRI) acquisitions, each differently sensitized to water diffusion. DTI computations are fundamentally estimation processes and are sensitive to noise and artifacts. Despite widespread adoption in the neuroimaging community, maintaining consistent DTI data quality remains challenging given the propensity for patient motion, artifacts associated with fast imaging techniques, and the possibility of hardware changes/failures. Furthermore, the quantity of data acquired per voxel, the non-linear estimation process, and numerous potential use cases complicate traditional visual data inspection approaches. Currently, quality inspection of DTI data has relied on visual inspection and individual processing in DTI analysis software programs (e.g. DTIPrep, DTI-studio). However, recent advances in applied statistical methods have yielded several different metrics to assess noise level, artifact propensity, quality of tensor fit, variance of estimated measures, and bias in estimated measures. To date, these metrics have been largely studied in isolation. Herein, we select complementary metrics for integration into an automatic DTI analysis and quality assurance pipeline. The pipeline completes in 24 hours, stores statistical outputs, and produces a graphical summary quality analysis (QA) report. We assess the utility of this streamlined approach for empirical quality assessment on 608 DTI datasets from pediatric neuroimaging studies. The efficiency and accuracy of quality analysis using the proposed pipeline is compared with quality analysis based on visual inspection. The unified pipeline is found to save a statistically significant amount of time (over 70%) while improving the consistency of QA between a DTI expert and a pool of research associates. Projection of QA metrics to a low dimensional manifold reveal qualitative, but clear, QA-study associations and suggest that automated outlier/anomaly detection would be feasible. PMID:23637895

  14. Image quality, space-qualified UV interference filters

    NASA Technical Reports Server (NTRS)

    Mooney, Thomas A.

    1992-01-01

    The progress during the contract period is described. The project involved fabrication of image quality, space-qualified bandpass filters in the 200-350 nm spectral region. Ion-assisted deposition (IAD) was applied to produce stable, reasonably durable filter coatings on space compatible UV substrates. Thin film materials and UV transmitting substrates were tested for resistance to simulated space effects.

  15. Perceived interest versus overt visual attention in image quality assessment

    NASA Astrophysics Data System (ADS)

    Engelke, Ulrich; Zhang, Wei; Le Callet, Patrick; Liu, Hantao

    2015-03-01

    We investigate the impact of overt visual attention and perceived interest on the prediction performance of image quality metrics. Towards this end we performed two respective experiments to capture these mechanisms: an eye gaze tracking experiment and a region-of-interest selection experiment. Perceptual relevance maps were created from both experiments and integrated into the design of the image quality metrics. Correlation analysis shows that indeed there is an added value of integrating these perceptual relevance maps. We reveal that the improvement in prediction accuracy is not statistically different between fixation density maps from eye gaze tracking data and region-of-interest maps, thus, indicating the robustness of different perceptual relevance maps for the performance gain of image quality metrics. Interestingly, however, we found that thresholding of region-of-interest maps into binary maps significantly deteriorates prediction performance gain for image quality metrics. We provide a detailed analysis and discussion of the results as well as the conceptual and methodological differences between capturing overt visual attention and perceived interest.

  16. SCID: full reference spatial color image quality metric

    NASA Astrophysics Data System (ADS)

    Ouni, S.; Chambah, M.; Herbin, M.; Zagrouba, E.

    2009-01-01

    The most used full reference image quality assessments are error-based methods. Thus, these measures are performed by pixel based difference metrics like Delta E ( E), MSE, PSNR, etc. Therefore, a local fidelity of the color is defined. However, these metrics does not correlate well with the perceived image quality. Indeed, they omit the properties of the HVS. Thus, they cannot be a reliable predictor of the perceived visual quality. All this metrics compute the differences pixel to pixel. Therefore, a local fidelity of the color is defined. However, the human visual system is rather sensitive to a global quality. In this paper, we present a novel full reference color metric that is based on characteristics of the human visual system by considering the notion of adjacency. This metric called SCID for Spatial Color Image Difference, is more perceptually correlated than other color differences such as Delta E. The suggested full reference metric is generic and independent of image distortion type. It can be used in different application such as: compression, restoration, etc.

  17. An image quality evaluation tool simulating image sensors including quantum efficiency off-axis effect

    NASA Astrophysics Data System (ADS)

    Mornet, Clémence; Vaillant, Jérôme; Decroux, Thomas; Virollet, Nicolas; Herault, Didier; Schanen, Isabelle

    2011-01-01

    The image quality evaluation of CMOS sensors is a big challenge for camera module manufacturers. In this paper, we present an update of the Image Quality Evaluation Tool, a graphical user interface simulating image sensors to assess the performance of a pixel. The simulated images are computed from operating conditions and sensor's characteristics like Quantum Efficiency including off-axis effect. Simulation of QE off-axis impact has been based on characterization data. The method does not require optics, making it suitable for early design phases as for optimizations and investigations. Both measurement and implementation in the tool will be explained. The QE degradation with angle effect will be highlighted on simulated images. A uniform gray scene or coloured image simulation from QE off-axis measurement will help engineers to calculate post-processing digital correction like colour shading correction or colour correction matrix versus pixel position.

  18. Comparison of quality control software tools for diffusion tensor imaging.

    PubMed

    Liu, Bilan; Zhu, Tong; Zhong, Jianhui

    2015-04-01

    Image quality of diffusion tensor imaging (DTI) is critical for image interpretation, diagnostic accuracy and efficiency. However, DTI is susceptible to numerous detrimental artifacts that may impair the reliability and validity of the obtained data. Although many quality control (QC) software tools are being developed and are widely used and each has its different tradeoffs, there is still no general agreement on an image quality control routine for DTIs, and the practical impact of these tradeoffs is not well studied. An objective comparison that identifies the pros and cons of each of the QC tools will be helpful for the users to make the best choice among tools for specific DTI applications. This study aims to quantitatively compare the effectiveness of three popular QC tools including DTI studio (Johns Hopkins University), DTIprep (University of North Carolina at Chapel Hill, University of Iowa and University of Utah) and TORTOISE (National Institute of Health). Both synthetic and in vivo human brain data were used to quantify adverse effects of major DTI artifacts to tensor calculation as well as the effectiveness of different QC tools in identifying and correcting these artifacts. The technical basis of each tool was discussed, and the ways in which particular techniques affect the output of each of the tools were analyzed. The different functions and I/O formats that three QC tools provide for building a general DTI processing pipeline and integration with other popular image processing tools were also discussed. PMID:25460331

  19. Pilot model of the extrahigh-quality imaging system

    NASA Astrophysics Data System (ADS)

    Taniho, Shuji; Ito, Hiroshi; Moriwaki, Hirohumi; Katada, Hideo; Makino, Shiro

    1996-02-01

    The progress of computer graphics and display technology has led us to always obtain an advanced visual image. However, we now feel the limit of the color reproduction (by the present 24 bits/pixel quantization, R, G, B, 8 bits respectively,) when pursuing a higher image quality. Therefore, we are developing an 'extra high quality imaging system' of 36 bit/pixel quantization (R, G, B, 12 bits, respectively.) This system comprises a MO disk drive, a controlling computer, a frame buffer and two 21' displays. The 2048 multiplied by 2048 pixel (36 bits/pixel) image data are read from the MO disk drive, and are sent to the frame buffer. A deliberately constructed 16 M byte frame buffer outputs the 36 bits/pixel video signal at a 200 MHz clock rate. Two displays, using a shadow-mask type CRT, are driven at a 78.7 kHz horizontal frequency. The system outputs the 36 bits/pixel and the 24 bits/pixel video signals concurrently, which makes it possible to compare the image quality of a 36 bits/pixel system with that of a 24 bits/pixel system. Many characteristics and physical factors, including noise, which do not cause a serious problem in conventional 24 bits/pixel systems, have a much more serious effect on the 36 bits/pixel system. We have now obtained the performance of color depth of up to 33 bits/pixel.

  20. MR image quality improvement by zero-filling Fresnel transform imaging technique

    NASA Astrophysics Data System (ADS)

    Wu, Bin Rong; Ito, Satoshi; Kamimura, Yoshitsugu; Yamada, Yoshifumi

    2005-04-01

    The quality of MR image is an important factor to control the accuracy of the diagnosis and treatment, so the image quality improvement methods, which can remove the noise and do not deteriorate the spatial resolution of images, are demanded in medical imaging field. There is an MR image reconstruction method, which applies the Fourier transform after a quadratic phase modulation in the Fresnel transform imaging technique. This method has a great property that can diffuse the noise contained in NMR signal by the quadratic phase modulation. If we prepare suitable area for diffusion of the noise by using zero-filling technique before reconstructing an image with this method, the spectrum of noise will evenly distribute over the zero filled area, and on the other hand, the image spectrum appear as an aspect of multi-resolution type image which locally distribute. After the noise was removed by a threshold filter in this aspect space, we return the data to the beginning signal space by doing a series of inverse processes, and reconstruct the denoised signal by using usual Fresnel transform image reconstruction technique. An MR image, which the noise is greatly improved, whereas the deterioration in spatial resolution is hardly caused, can be obtained. In this work, we present a new MR image quality improvement method that uses the property of noise diffusing in the Fresnel transform imaging technique; and we describe the effectiveness from simulations that are evaluated on SNR improvement and extent of deterioration in spatial resolution, and compare them with standard Wiener filtering and Wavelet Wiener filtering. Finally, we verify the powerfulness of the proposed method and it is applicable to phase-scrambled Fourier transform imaging technique through experiment.

  1. Comparison of the polynomial model against explicit measurements of noise components for different mammography systems.

    PubMed

    Monnin, P; Bosmans, H; Verdun, F R; Marshall, N W

    2014-10-01

    Given the adverse impact of image noise on the perception of important clinical details in digital mammography, routine quality control measurements should include an evaluation of noise. The European Guidelines, for example, employ a second-order polynomial fit of pixel variance as a function of detector air kerma (DAK) to decompose noise into quantum, electronic and fixed pattern (FP) components and assess the DAK range where quantum noise dominates. This work examines the robustness of the polynomial method against an explicit noise decomposition method. The two methods were applied to variance and noise power spectrum (NPS) data from six digital mammography units. Twenty homogeneously exposed