Science.gov

Sample records for pet scanner clairvivopet

  1. Performance evaluation of the small-animal PET scanner ClairvivoPET using NEMA NU 4-2008 Standards

    NASA Astrophysics Data System (ADS)

    Sato, K.; Shidahara, M.; Watabe, H.; Watanuki, S.; Ishikawa, Y.; Arakawa, Y.; Nai, YH; Furumoto, S.; Tashiro, M.; Shoji, T.; Yanai, K.; Gonda, K.

    2016-01-01

    The aim of this study was to evaluate the performance of ClairvivoPET using NEMA NU4 standards. The ClairvivoPET incorporates a LYSO dual depth-of-interaction detector system with 151 mm axial field of view (FOV). Spatial resolution, sensitivity, counting rate capabilities, and image quality were evaluated using NEMA NU4-2008 standards. Normal mouse imaging was also performed for 10min after intravenous injection of 18F(-)-NaF. Data were compared with 19 other preclinical PET scanners. Spatial resolution measured using full width at half maximum on FBP-ramp reconstructed images was 2.16 mm at radial offset 5 mm of the axial centre FOV. The maximum absolute sensitivity for a point source at the FOV centre was 8.72%. Peak noise equivalent counting rate (NECR) was 415kcps at 14.6MBq ml-1. The uniformity with the image-quality phantom was 4.62%. Spillover ratios in the images of air and water filled chambers were 0.19 and 0.06, respectively. Our results were comparable with the 19 other preclinical PET scanners based on NEMA NU4 standards, with excellent sensitivity because of the large FOV. The ClairvivoPET with iterative reconstruction algorithm also provided sufficient visualization of the mouse spine. The high sensitivity and resolution of the ClairvivoPET scanner provided high quality images for preclinical studies.

  2. Performance evaluation of the small-animal PET scanner ClairvivoPET using NEMA NU 4-2008 Standards.

    PubMed

    Sato, K; Shidahara, M; Watabe, H; Watanuki, S; Ishikawa, Y; Arakawa, Y; Nai, Y H; Furumoto, S; Tashiro, M; Shoji, T; Yanai, K; Gonda, K

    2016-01-21

    The aim of this study was to evaluate the performance of ClairvivoPET using NEMA NU4 standards. The ClairvivoPET incorporates a LYSO dual depth-of-interaction detector system with 151 mm axial field of view (FOV). Spatial resolution, sensitivity, counting rate capabilities, and image quality were evaluated using NEMA NU4-2008 standards. Normal mouse imaging was also performed for 10 min after intravenous injection of (18)F(-)-NaF. Data were compared with 19 other preclinical PET scanners. Spatial resolution measured using full width at half maximum on FBP-ramp reconstructed images was 2.16 mm at radial offset 5 mm of the axial centre FOV. The maximum absolute sensitivity for a point source at the FOV centre was 8.72%. Peak noise equivalent counting rate (NECR) was 415 kcps at 14.6 MBq ml(-1). The uniformity with the image-quality phantom was 4.62%. Spillover ratios in the images of air and water filled chambers were 0.19 and 0.06, respectively. Our results were comparable with the 19 other preclinical PET scanners based on NEMA NU4 standards, with excellent sensitivity because of the large FOV. The ClairvivoPET with iterative reconstruction algorithm also provided sufficient visualization of the mouse spine. The high sensitivity and resolution of the ClairvivoPET scanner provided high quality images for preclinical studies. PMID:26716872

  3. Combined PET/MRI scanner

    DOEpatents

    Schlyer, David; Woody, Craig L.; Rooney, William; Vaska, Paul; Stoll, Sean; Pratte, Jean-Francois; O'Connor, Paul

    2007-10-23

    A combined PET/MRI scanner generally includes a magnet for producing a magnetic field suitable for magnetic resonance imaging, a radiofrequency (RF) coil disposed within the magnetic field produced by the magnet and a ring tomograph disposed within the magnetic field produced by the magnet. The ring tomograph includes a scintillator layer for outputting at least one photon in response to an annihilation event, a detection array coupled to the scintillator layer for detecting the at least one photon outputted by the scintillator layer and for outputting a detection signal in response to the detected photon and a front-end electronic array coupled to the detection array for receiving the detection signal, wherein the front-end array has a preamplifier and a shaper network for conditioning the detection signal.

  4. Miniature 'Wearable' PET Scanner Ready for Use

    SciTech Connect

    Paul Vaska

    2011-03-09

    Scientists from BNL, Stony Brook University, and collaborators have demonstrated the efficacy of a "wearable," portable PET scanner they've developed for rats. The device will give neuroscientists a new tool for simultaneously studying brain function and behavior in fully awake, moving animals.

  5. Miniature 'Wearable' PET Scanner Ready for Use

    ScienceCinema

    Paul Vaska

    2013-07-22

    Scientists from BNL, Stony Brook University, and collaborators have demonstrated the efficacy of a "wearable," portable PET scanner they've developed for rats. The device will give neuroscientists a new tool for simultaneously studying brain function and behavior in fully awake, moving animals.

  6. Development of scintillation materials for PET scanners

    NASA Astrophysics Data System (ADS)

    Korzhik, Mikhail; Fedorov, Andrei; Annenkov, Alexander; Borissevitch, Andrei; Dossovitski, Alexei; Missevitch, Oleg; Lecoq, Paul

    2007-02-01

    The growing demand on PET methodology for a variety of applications ranging from clinical use to fundamental studies triggers research and development of PET scanners providing better spatial resolution and sensitivity. These efforts are primarily focused on the development of advanced PET detector solutions and on the developments of new scintillation materials as well. However Lu containing scintillation materials introduced in the last century such as LSO, LYSO, LuAP, LuYAP crystals still remain the best PET species in spite of the recent developments of bright, fast but relatively low density lanthanum bromide scintillators. At the same time Lu based materials have several drawbacks which are high temperature of crystallization and relatively high cost compared to alkali-halide scintillation materials. Here we describe recent results in the development of new scintillation materials for PET application.

  7. Design and performance of HEAD PENN-PET scanner

    SciTech Connect

    Freifelder, R.; Karp, J.S. . Dept. of Radiology); Geagan, M.; Muehllehner, G. )

    1994-08-01

    A new PET scanner for brain imaging (and animals) has been designed with very high sensitivity and spatial resolution. The design is an evolution of the PENN-PET scanner, which uses large position-sensitive NaI(Tl) detectors, with Anger-type positioning logic, and which allows 3-D volume imaging, without septa. The new design is built with a single annular crystal coupled to 180 photomultiplier tubes, and uses local triggering electronics to subdivide the detector into small zones and to determine coincident events within the detector. The axial acceptance angle of [+-] 27 deg, with a field-of-view of 25.6 cm, is larger than any currently operating PET scanner. Performance measurements are presented.

  8. Temperature dependence of APD-based PET scanners

    SciTech Connect

    Keereman, Vincent; Van Holen, Roel; Vandenberghe, Stefaan; Vanhove, Christian

    2013-09-15

    Purpose: Solid state detectors such as avalanche photodiodes (APDs) are increasingly being used in PET detectors. One of the disadvantages of APDs is the strong decrease of their gain factor with increasing ambient temperature. The light yield of most scintillation crystals also decreases when ambient temperature is increased. Both effects lead to considerable temperature dependence of the performance of APD-based PET scanners. In this paper, the authors propose a model for this dependence and the performance of the LabPET8 APD-based small animal PET scanner is evaluated at different temperatures.Methods: The model proposes that the effect of increasing temperature on the energy histogram of an APD-based PET scanner is a compression of the histogram along the energy axis. The energy histogram of the LabPET system was acquired at 21 °C and 25 °C to verify the validity of this model. Using the proposed model, the effect of temperature on system sensitivity was simulated for different detector temperature coefficients and temperatures. Subsequently, the effect of short term and long term temperature changes on the peak sensitivity of the LabPET system was measured. The axial sensitivity profile was measured at 21 °C and 24 °C following the NEMA NU 4-2008 standard. System spatial resolution was also evaluated. Furthermore, scatter fraction, count losses and random coincidences were evaluated at different temperatures. Image quality was also investigated.Results: As predicted by the model, the photopeak energy at 25 °C is lower than at 21 °C with a shift of approximately 6% per °C. Simulations showed that this results in an approximately linear decrease of sensitivity when temperature is increased from 21 °C to 24 °C and energy thresholds are constant. Experimental evaluation of the peak sensitivity at different temperatures showed a strong linear correlation for short term (2.32 kcps/MBq/°C = 12%/°C, R = −0.95) and long term (1.92 kcps/MBq/°C = 10%/

  9. Development of a PET Scanner for Simultaneously Imaging Small Animals with MRI and PET

    PubMed Central

    Thompson, Christopher J; Goertzen, Andrew L; Thiessen, Jonathan D; Bishop, Daryl; Stortz, Greg; Kozlowski, Piotr; Retière, Fabrice; Zhang, Xuezhu; Sossi, Vesna

    2014-01-01

    Recently, positron emission tomography (PET) is playing an increasingly important role in the diagnosis and staging of cancer. Combined PET and X-ray computed tomography (PET-CT) scanners are now the modality of choice in cancer treatment planning. More recently, the combination of PET and magnetic resonance imaging (MRI) is being explored in many sites. Combining PET and MRI has presented many challenges since the photo-multiplier tubes (PMT) in PET do not function in high magnetic fields, and conventional PET detectors distort MRI images. Solid state light sensors like avalanche photo-diodes (APDs) and more recently silicon photo-multipliers (SiPMs) are much less sensitive to magnetic fields thus easing the compatibility issues. This paper presents the results of a group of Canadian scientists who are developing a PET detector ring which fits inside a high field small animal MRI scanner with the goal of providing simultaneous PET and MRI images of small rodents used in pre-clinical medical research. We discuss the evolution of both the crystal blocks (which detect annihilation photons from positron decay) and the SiPM array performance in the last four years which together combine to deliver significant system performance in terms of speed, energy and timing resolution. PMID:25120157

  10. Development of a PET scanner for simultaneously imaging small animals with MRI and PET.

    PubMed

    Thompson, Christopher J; Goertzen, Andrew L; Thiessen, Jonathan D; Bishop, Daryl; Stortz, Greg; Kozlowski, Piotr; Retière, Fabrice; Zhang, Xuezhu; Sossi, Vesna

    2014-01-01

    Recently, positron emission tomography (PET) is playing an increasingly important role in the diagnosis and staging of cancer. Combined PET and X-ray computed tomography (PET-CT) scanners are now the modality of choice in cancer treatment planning. More recently, the combination of PET and magnetic resonance imaging (MRI) is being explored in many sites. Combining PET and MRI has presented many challenges since the photo-multiplier tubes (PMT) in PET do not function in high magnetic fields, and conventional PET detectors distort MRI images. Solid state light sensors like avalanche photo-diodes (APDs) and more recently silicon photo-multipliers (SiPMs) are much less sensitive to magnetic fields thus easing the compatibility issues. This paper presents the results of a group of Canadian scientists who are developing a PET detector ring which fits inside a high field small animal MRI scanner with the goal of providing simultaneous PET and MRI images of small rodents used in pre-clinical medical research. We discuss the evolution of both the crystal blocks (which detect annihilation photons from positron decay) and the SiPM array performance in the last four years which together combine to deliver significant system performance in terms of speed, energy and timing resolution. PMID:25120157

  11. MR-assisted PET Motion Correction for eurological Studies in an Integrated MR-PET Scanner

    PubMed Central

    Catana, Ciprian; Benner, Thomas; van der Kouwe, Andre; Byars, Larry; Hamm, Michael; Chonde, Daniel B.; Michel, Christian J.; El Fakhri, Georges; Schmand, Matthias; Sorensen, A. Gregory

    2011-01-01

    Head motion is difficult to avoid in long PET studies, degrading the image quality and offsetting the benefit of using a high-resolution scanner. As a potential solution in an integrated MR-PET scanner, the simultaneously acquired MR data can be used for motion tracking. In this work, a novel data processing and rigid-body motion correction (MC) algorithm for the MR-compatible BrainPET prototype scanner is described and proof-of-principle phantom and human studies are presented. Methods To account for motion, the PET prompts and randoms coincidences as well as the sensitivity data are processed in the line or response (LOR) space according to the MR-derived motion estimates. After sinogram space rebinning, the corrected data are summed and the motion corrected PET volume is reconstructed from these sinograms and the attenuation and scatter sinograms in the reference position. The accuracy of the MC algorithm was first tested using a Hoffman phantom. Next, human volunteer studies were performed and motion estimates were obtained using two high temporal resolution MR-based motion tracking techniques. Results After accounting for the physical mismatch between the two scanners, perfectly co-registered MR and PET volumes are reproducibly obtained. The MR output gates inserted in to the PET list-mode allow the temporal correlation of the two data sets within 0.2 s. The Hoffman phantom volume reconstructed processing the PET data in the LOR space was similar to the one obtained processing the data using the standard methods and applying the MC in the image space, demonstrating the quantitative accuracy of the novel MC algorithm. In human volunteer studies, motion estimates were obtained from echo planar imaging and cloverleaf navigator sequences every 3 seconds and 20 ms, respectively. Substantially improved PET images with excellent delineation of specific brain structures were obtained after applying the MC using these MR-based estimates. Conclusion A novel MR-based MC

  12. Monte Carlo simulation of efficient data acquisition for an entire-body PET scanner

    NASA Astrophysics Data System (ADS)

    Isnaini, Ismet; Obi, Takashi; Yoshida, Eiji; Yamaya, Taiga

    2014-07-01

    Conventional PET scanners can image the whole body using many bed positions. On the other hand, an entire-body PET scanner with an extended axial FOV, which can trace whole-body uptake images at the same time and improve sensitivity dynamically, has been desired. The entire-body PET scanner would have to process a large amount of data effectively. As a result, the entire-body PET scanner has high dead time at a multiplex detector grouping process. Also, the entire-body PET scanner has many oblique line-of-responses. In this work, we study an efficient data acquisition for the entire-body PET scanner using the Monte Carlo simulation. The simulated entire-body PET scanner based on depth-of-interaction detectors has a 2016-mm axial field-of-view (FOV) and an 80-cm ring diameter. Since the entire-body PET scanner has higher single data loss than a conventional PET scanner at grouping circuits, the NECR of the entire-body PET scanner decreases. But, single data loss is mitigated by separating the axially arranged detector into multiple parts. Our choice of 3 groups of axially-arranged detectors has shown to increase the peak NECR by 41%. An appropriate choice of maximum ring difference (MRD) will also maintain the same high performance of sensitivity and high peak NECR while at the same time reduces the data size. The extremely-oblique line of response for large axial FOV does not contribute much to the performance of the scanner. The total sensitivity with full MRD increased only 15% than that with about half MRD. The peak NECR was saturated at about half MRD. The entire-body PET scanner promises to provide a large axial FOV and to have sufficient performance values without using the full data.

  13. Iterative reconstruction for pet scanners with continuous scintillators.

    PubMed

    Iriarte, Ana; Caffarena, Gabriel; Lopez-Fernandez, Mariano; Garcia-Carmona, Rodrigo; Otero, Abraham; Sorzano, Carlos O S; Marabini, Roberto

    2015-08-01

    Several technical developments have led to a comeback of the continuous scintillators in positron emission tomography (PET). Important differences exist between the resurgent continuous scintillators and the prevailing pixelated devices, which can translate into certain advantages of the former over the latter. However, if the peculiarities of the continuous scintillators are not considered in the iterative reconstruction in which the measured data is converted to images, these advantages will not be fully exploited. In this paper, we review which those peculiarities are and how they have been considered in the literature of PET reconstruction. In light of this review, we propose a new method to compute one of the key elements of the iterative schemes, the system matrix. Specifically, we substitute the traditional Gaussian approach to the so-called uncertainty term by a more general Monte Carlo estimation, and account for the effect of the optical photons, which cannot be neglected in continuous-scintillators devices. Finally, we gather in a single scheme all the elements of the iterative reconstruction that have been individually reformulated, in this or previous works, for continuous scintillators, providing the first reconstruction framework fully adapted to this type of detectors. The preliminary images obtained for a commercially available PET scanner show the benefits of adjusting the reconstruction to the nature of the scintillators. PMID:26736742

  14. Performance evaluation and calibration of the neuro-pet scanner

    SciTech Connect

    Sank, V.J.; Brooks, R.A.; Cascio, H.E.; Di Chiro, G.; Friauf, W.S.; Leighton, S.B.

    1983-02-01

    The Neuro-PET is a circular ring seven-slice positron emission tomograph designed for imaging human heads and small animals. The scanner uses 512 bismuth germanate detectors 8.25 mm wide packed tightly together in four layers to achieve high spatial resolution (6-7 mm FWHM) without the use of beam blockers. Because of the small 38 cm ring diameter, the sensitivity is also very high: 70,000 c/s per true slice with medium energy threshold (375 keV) for a 20 cm diameter phantom containing 1 ..mu..Ci/cc of positron-emitting activity, according to a preliminary measurement. There are three switch-selectable thresholds, and the sensitivity will be higher in the low threshold setting. The Neuro-PET is calibrated with a round or elliptical phantom that approximates a patient's head; this method eliminates the effects of scatter and self-attenuation to first order. Further software corrections for these artifacts are made in the reconstruction program, which reduce the measured scatter to zero, as determined with a 5 cm cold spot. With a 1 cm cold spot, the apparent activity at the center of the cold spot is 18% of the surrounding activity, which is clearly a consequence of the limits of spatial resolution, rather than scatter. The Neuro-PET has been in clinical operation since June 1982, and approximately 30 patients have been scanned to date.

  15. Efficient system modeling for a small animal PET scanner with tapered DOI detectors

    NASA Astrophysics Data System (ADS)

    Zhang, Mengxi; Zhou, Jian; Yang, Yongfeng; Rodríguez-Villafuerte, Mercedes; Qi, Jinyi

    2016-01-01

    A prototype small animal positron emission tomography (PET) scanner for mouse brain imaging has been developed at UC Davis. The new scanner uses tapered detector arrays with depth of interaction (DOI) measurement. In this paper, we present an efficient system model for the tapered PET scanner using matrix factorization and a virtual scanner geometry. The factored system matrix mainly consists of two components: a sinogram blurring matrix and a geometrical matrix. The geometric matrix is based on a virtual scanner geometry. The sinogram blurring matrix is estimated by matrix factorization. We investigate the performance of different virtual scanner geometries. Both simulation study and real data experiments are performed in the fully 3D mode to study the image quality under different system models. The results indicate that the proposed matrix factorization can maintain image quality while substantially reduce the image reconstruction time and system matrix storage cost. The proposed method can be also applied to other PET scanners with DOI measurement.

  16. Performance evaluation of an Inveon PET preclinical scanner.

    PubMed

    Constantinescu, Cristian C; Mukherjee, Jogeshwar

    2009-05-01

    We evaluated the performance of an Inveon preclinical PET scanner (Siemens Medical Solutions), the latest MicroPET system. Spatial resolution was measured with a glass capillary tube (0.26 mm inside diameter, 0.29 mm wall thickness) filled with (18)F solution. Transaxial and axial resolutions were measured with the source placed parallel and perpendicular to the axis of the scanner. The sensitivity of the scanner was measured with a (22)Na point source, placed on the animal bed and positioned at different offsets from the center of the field of view (FOV), as well as at different energy and coincidence windows. The noise equivalent count rates (NECR) and the system scatter fraction were measured using rat-like (Phi = 60, L = 150 mm) and mouse-like (Phi = 25 mm, L = 70 mm) cylindrical phantoms. Line sources filled with high activity (18)F (>250 MBq) were inserted parallel to the axes of the phantoms (13.5 and 10 mm offset). For each phantom, list-mode data were collected over 24 h at 350-650 keV and 250-750 keV energy windows and 3.4 ns coincidence window. System scatter fraction was measured when the random event rates were below 1%. Performance phantoms consisting of cylinders with hot rod inserts filled with (18)F were imaged. In addition, we performed imaging studies that show the suitability of the Inveon scanner for imaging small structures such as those in mice with a variety of tracers. The radial, tangential and axial resolutions at the center of FOV were 1.46 mm, 1.49 and 1.15 mm, respectively. At a radial offset of 2 cm, the FWHM values were 1.73, 2.20 and 1.47 mm, respectively. At a coincidence window of 3.4 ns, the sensitivity was 5.75% for EW = 350-650 keV and 7.4% for EW = 250-750 keV. For an energy window of 350-650 keV, the peak NECR was 538 kcps at 131.4 MBq for the rat-like phantom, and 1734 kcps at 147.4 MBq for the mouse-like phantom. The system scatter fraction values were 0.22 for the rat phantom and 0.06 for the mouse phantom. The Inveon

  17. Hybrid approach for attenuation correction in PET/MR scanners

    NASA Astrophysics Data System (ADS)

    Santos Ribeiro, A.; Rota Kops, E.; Herzog, H.; Almeida, P.

    2014-01-01

    Aim: Attenuation correction (AC) of PET images is still one of the major limitations of hybrid PET/MR scanners. Different methods have been proposed to obtain the AC map from morphological MR images. Although, segmentation methods normally fail to differentiate air and bone regions, while template or atlas methods usually cannot accurately represent regions anatomically different from the template image. In this study a feed forward neural network (FFNN) algorithm is presented which directly outputs the attenuation coefficients by non-linear regression of the images acquired with an ultrashort echo time (UTE) sequence guided by the template-based AC map (TAC-map). Materials and methods: MR as well as CT data were acquired in four subjects. The UTE images and the TAC-map were the inputs of the presented FFNN algorithm for training as well as classification. The resulting attenuation maps were compared with CT-based, PNN-based and TAC maps. All the AC maps were used to reconstruct the PET emission data which were then compared for the different methods. Results: For each subject dice coefficients D were calculated between each method and the respective CT-based AC maps. The resulting Ds show higher values for all FFNN-based tissues comparatively to both TAC-based and PNN-based methods, particularly for bone tissue (D=0.77, D=0.51 and D=0.71, respectively). The AC-corrected PET images with the FFNN-based map show an overall lower relative difference (RD=3.90%) than those AC-corrected with the PNN-based (RD=4.44%) or template-based (RD=4.43%) methods. Conclusion: Our results show that an enhancement of current methods can be performed by combining both information of new MR image sequence techniques and general information provided from template techniques. Nevertheless, the number of tested subjects is statistically low and current analysis for a larger dataset is being carried out.

  18. Data Acquisition and Image Reconstruction Systems from the miniPET Scanners to the CARDIOTOM Camera

    SciTech Connect

    Valastvan, I.; Imrek, J.; Hegyesi, G.; Molnar, J.; Novak, D.; Bone, D.; Kerek, A.

    2007-11-26

    Nuclear imaging devices play an important role in medical diagnosis as well as drug research. The first and second generation data acquisition systems and the image reconstruction library developed provide a unified hardware and software platform for the miniPET-I, miniPET-II small animal PET scanners and for the CARDIOTOM{sup TM}.

  19. Improved dead-time correction for PET scanners: application to small-animal PET

    NASA Astrophysics Data System (ADS)

    Vicente, E.; Herraiz, J. L.; España, S.; Herranz, E.; Desco, M.; Vaquero, J. J.; Udías, J. M.

    2013-04-01

    Pile-up and dead-time are two main causes of nonlinearity in the response of a PET scanner as a function of activity in the field of view (FOV). For a given scanner and acquisition system, pile-up effects depend on the material and size of the object being imaged and on the distribution of activity inside and outside the FOV, because these factors change the singles-to-coincidences ratio (SCR). Thus, it is difficult to devise an accurate correction that would be valid for any acquisition. In this work, we demonstrate a linear relationship between SCR and effective dead-time, which measures the effects of both dead-time (losses) and pile-up (gains and losses). This relationship allows us to propose a simple method to accurately estimate dead-time and pile-up corrections using only two calibration acquisitions with, respectively, a high and low SCR. The method has been tested with simulations and experimental data for two different scanner geometries: a scanner with large area detectors and no pile-up rejection, and a scanner composed of two full rings of smaller detectors. Our results show that the SCR correction method is accurate within 7%, even for high activities in the FOV, and avoids the bias of the standard single-parameter method.

  20. Feasibility study of small animal imaging using clinical PET/CT scanner

    NASA Astrophysics Data System (ADS)

    Hsu, Wen-Lin; Chen, Chia-Lin; Wang, Ze-Jing; Wu, Tung-Hsin; Liu, Dai-Wei; Lee, Jason J. S.

    2007-02-01

    The feasibility of small animal imaging using a clinical positron emission tomography/computed tomography (PET/CT) scanner with [F-18]-fluoro-2-deoxy- D-glucose (FDG) was evaluated. Two protocols in PET/CT system, single-mouse high-resolution mode (SHR) and multi-mouse high throughput mode (MHT) protocol were employed to investigate the ability of the scanner and also explored the performance differences between microPET and clinical PET/CT. In this study, we have found that even the clinical PET/CT scanner could not compete with the microPET scanner, especially in spatial resolution; the high-resolution CT image could advance the anatomical information to sub-millimeter level. Besides, CT-based attenuation correction can improve the image uniformity characteristics and quantification accuracy, and the large bore of a human whole-body scanner broadens the possibility of high throughput studies. Considering all the benefits, clinical PET/CT imaging might be a potential alternative for small animal study.

  1. Study of PET scanner designs using clinical metrics to optimize the scanner axial FOV and crystal thickness

    NASA Astrophysics Data System (ADS)

    Surti, S.; Werner, M. E.; Karp, J. S.

    2013-06-01

    The aim of this study is to understand the trade-off between crystal thickness and scanner axial field-of-view FOV (AFOV) for clinical PET imaging. Clinical scanner design has evolved towards 20-25 mm thick crystals and 16-22 cm long scanner AFOV, as well as time-of-flight (TOF) imaging. While Monte Carlo studies demonstrate that longer AFOV and thicker crystals will lead to higher scanner sensitivity, cost has prohibited the building of commercial scanners with >22 cm AFOV. In this study, we performed a series of system simulations to optimize the use of a given amount of crystal material by evaluating the impact on system sensitivity and noise equivalent counts (NEC), as well as image quality in terms of lesion detectability. We evaluated two crystal types (LSO and LaBr3) and fixed the total crystal volume used for each type (8.2 L of LSO and 17.1 L of LaBr3) while varying the crystal thickness and scanner AFOV. In addition, all imaging times were normalized so that the total scan time needed to scan a 100 cm long object with multiple bed positions was kept constant. Our results show that the highest NEC cm-1 in a 35 cm diameter ×70 cm long line source cylinder is achieved for an LSO scanner with 10 mm long crystals and AFOV of 36 cm, while for LaBr3 scanners, the highest NEC cm-1 is obtained with 20 mm long crystals and an AFOV of 38 cm. Lesion phantom simulations show that the best lesion detection performance is achieved in scanners with long AFOV (≥36 cm) and using thin crystals (≤10 mm of LSO and ≤20 mm of LaBr3). This is due to a combination of improved NEC, as well as improved lesion contrast estimation due to better spatial resolution in thinner crystals. Alternatively, for lesion detection performance similar to that achieved in standard clinical scanner designs, the long AFOV scanners can be used to reduce the total scan time without increasing the amount of crystal used in the scanner. In addition, for LaBr3 based scanners, the reduced lesion

  2. Development of prototype PET scanner using dual-sided readout DOI-PET modules

    NASA Astrophysics Data System (ADS)

    Fujita, T.; Kataoka, J.; Kishimoto, A.; Kurei, Y.; Nishiyama, T.; Ohshima, T.; Taya, T.

    2014-12-01

    In our previous work, we proposed a novel design for a gamma-ray detector module capable of measuring the depth of interaction (DOI). In this paper, we further developed DOI-PET detector modules and a data acquisition system, and evaluated its performance. Each detector module was composed of a 3-D scintillator array and two large-area monolithic Multi-Pixel Photon Counter (MPPC) arrays coupled to both ends of the 3-D scintillator array, leading to only 8-ch signal outputs from a module. The 3-D scintillator array was composed of 9 × 9 × 7 matrices of 1.0 × 1.0 × 3.0 mm3 Ce:GAGG crystals. The detector module showed good energy resolution of 10.6% as measured at 511 keV and a high average peak to valley ratio higher than 8 for each pixel crystal identified in the X-Y-Z directions. Moreover, we evaluated the spatial resolution of a virtual 18-ch PET gantry simulated by using two detector modules that were flexibly controlled using both the X-stage and θ-stage in 20-degree steps. By measuring a 22Na point source (0.25 mm in diameter), we showed that spatial resolution substantially degrades from 1.0 mm to 7.8 mm (FWHM; as measured at 0 mm and 28 mm off-center) with a non-DOI configuration, whereas the corresponding values for the DOI configuration were 0.9 mm and 1.5 mm, respectively (FWHM; as measured at 0 mm and 28 mm off-center). This preliminary study confirms that our DOI-PET module is useful for future high spatial resolution and compact small-animal PET scanners without radial image distortions at the peripheral regions of the field of view (FOV).

  3. Voxel-based classification of FDG PET in dementia using inter-scanner normalization.

    PubMed

    Thiele, Frank; Young, Stewart; Buchert, Ralph; Wenzel, Fabian

    2013-08-15

    Statistical mapping of FDG PET brain images has become a common tool in differential diagnosis of patients with dementia. We present a voxel-based classification system of neurodegenerative dementias based on partial least squares (PLS). Such a classifier relies on image databases of normal controls and dementia cases as training data. Variations in PET image characteristics can be expected between databases, for example due to differences in instrumentation, patient preparation, and image reconstruction. This study evaluates (i) the impact of databases from different scanners on classification accuracy and (ii) a method to improve inter-scanner classification. Brain FDG PET databases from three scanners (A, B, C) at two clinical sites were evaluated. Diagnostic categories included normal controls (NC, nA=26, nB=20, nC=24 for each scanner respectively), Alzheimer's disease (AD, nA=44, nB=11, nC=16), and frontotemporal dementia (FTD, nA=13, nB=13, nC=5). Spatially normalized images were classified as NC, AD, or FTD using partial least squares. Supervised learning was employed to determine classifier parameters, whereby available data is sub-divided into training and test sets. Four different database setups were evaluated: (i) "in-scanner": training and test data from the same scanner, (ii) "x-scanner": training and test data from different scanners, (iii) "train other": train on both x-scanners, and (iv) "train all": train on all scanners. In order to moderate the impact of inter-scanner variations on image evaluation, voxel-by-voxel scaling was applied based on "ratio images". Good classification accuracy of on average 94% was achieved for the in-scanner setups. Accuracy deteriorated for setups with mismatched scanners (79-91%). Ratio-image normalization improved all results with mismatched scanners (85-92%). In conclusion, automatic classification of individual FDG PET in differential diagnosis of dementia is feasible. Accuracy can vary with respect to scanner or

  4. Preventive maintenance system for the photomultiplier detector blocks of PET scanners

    DOEpatents

    Levy, A.V.; Warner, D.

    1995-01-24

    A system including a method and apparatus for preventive maintenance of PET scanner photomultiplier detector blocks is disclosed. The qualitative comparisons used in the method of the present invention to provide an indication in the form of a display or printout advising the user that the photomultiplier block is stable, intermittently unstable, or drifting unstable, and also advising of the expected date of failure of a photomultiplier block in the PET scanner. The system alerts the user to replace the defective photomultiplier block prior to catastrophic failure in a scheduled preventative maintenance program, thus eliminating expensive and unscheduled downtime of the PET scanner due to photomultiplier failure. The apparatus for carrying out the method of the present invention preferably resides in the host computer controlling a PET scanner. It includes a memory adapted for storing a record of a number of iterative adjustments that are necessary to calibrate the gain of a photomultiplier detector block i at a time t[sub 0], a time t[sub 1] and a time T, where T>t[sub 1]>t[sub 0], which is designated as Histo(i,j(t)). The apparatus also includes a processor configured by a software program or a combination of programmed RAM and ROM devices to perform a number of calculations and operations on these values, and also includes a counter for analyzing each photomultiplier detector block i=1 through I of a PET scanner. 40 figures.

  5. Preventive maintenance system for the photomultiplier detector blocks of pet scanners

    DOEpatents

    Levy, Alejandro V.; Warner, Donald

    1995-01-24

    A system including a method and apparatus for preventive maintenance of PET scanner photomultiplier detector blocks is disclosed. The quantitive comparisons used in the method of the present invention to provide an indication in the form of a display or printout advising the user that the photomultiplier block is stable, intermittently unstable, or drifting unstable, and also advising of the expected date of failure of a photomultiplier block in the PET scanner. The system alerts the user to replace the defective photomultiplier block prior to catastrophic failure in a scheduled preventative maintenance program, thus eliminating expensive and unscheduled downtime of the PET scanner due to photomultiplier failure. The apparatus for carrying out the method of the present invention preferably resides in the host computer controlling a PET scanner. It includes a memory adapted for storing a record of a number of iterative adjustments that are necessary to calibrate the gain of a photomultiplier detector block i at a time t.sub.0, a time t.sub.1 and a time T, where T>t.sub.1 >t.sub.0, which is designated as Histo(i,j(t)). The apparatus also includes a processor configured by a software program or a combination of programmed RAM and ROM devices to perform a number of calculations and operations on these values, and also includes a counter for analyzing each photomultiplier detector block i=1 through I of a PET scanner.

  6. Performance evaluation of a LYSO-based PET scanner for monitoring of dose delivery in hadrontherapy

    NASA Astrophysics Data System (ADS)

    Fabbiani, E.; Belcari, N.; Camarlinghi, N.; Del Guerra, A.; Ferretti, S.; Kraan, A.; Panetta, D.; Sportelli, G.; Rosso, V.

    2015-12-01

    The DoPET scanner is a compact positron emission tomography (PET) device. It has been developed for monitoring the range of charged particles during therapy with hadron beams. Previous works have focused on the development and upgrade of the device and on data analysis. In this paper, a full performance characterization of the DoPET system in terms of the energy resolution, spatial resolution, sensitivity, uniformity, and noise equivalent count rate is reported. All measurements refer to an adapted version of the National Electrical Manufacturers Association (NEMA) NU 4 - 2008 protocol, which was written originally for small animal PET systems. Since DoPET is a dual head planar system, it requires a modified characterisation procedure with respect to those described for ring geometries as in the NEMA NU 4 - 2008 protocol. The presented procedure may be of interest for any other PET system with a similar geometry as DoPET.

  7. Performance and Limitations of Positron Emission Tomography (PET) Scanners for Imaging Very Low Activity Sources

    PubMed Central

    Freedenberg, Melissa; Badawi, Ramsey D.; Tarantal, Alice F.; Cherry, Simon R.

    2013-01-01

    Emerging applications for positron emission tomography (PET) may require the ability to image very low activity source distributions in the body. The performance of clinical PET scanners in the regime where activity in the field of view is <1 MBq has not previously been explored. In this study, we compared the counting rate performance of two clinical PET/CT scanners, the Siemens Biograph Reveal 16 scanner which is based on lutetium oxyorthosilicate (LSO) detectors and the GE Discovery-ST scanner which is based on bismuth germanate (BGO) detectors using a modified National Electrical Manufacturers Association (NEMA) NU 2-2007 protocol. Across the activity range studied (2-100 kBq/mL in a 5.5 mL line source in the NEMA scatter phantom), the BGO-based scanner significantly outperformed the LSO-based scanner. This was largely due to the effect of background counts emanating from naturally occurring but radioactive 176Lu within the LSO detector material, which dominates the observed counting rate at the lowest activities. Increasing the lower energy threshold from 350 keV to 425 keV in an attempt to reduce this background did not significantly improve the measured NECR performance. The measured singles rate due to 176Lu emissions within the scanner energy window was also found to be dependent on temperature, and to be affected by the operation of CT component, making approaches to correct or compensate for the background more challenging. We conclude that for PET studies in a very low activity range, BGO-based scanners are likely to have better performance because of the lack of significant background. PMID:23680361

  8. The simulation of a data acquisition system for a proposed high resolution PET scanner

    SciTech Connect

    Rotolo, C.; Larwill, M.; Chappa, S.; Ordonez, C.

    1993-10-01

    The simulation of a specific data acquisition (DAQ) system architecture for a proposed high resolution Positron Emission Tomography (PET) scanner is discussed. Stochastic processes are used extensively to model PET scanner signal timing and probable DAQ circuit limitations. Certain architectural parameters, along with stochastic parameters, are varied to quantatively study the resulting output under various conditions. The inclusion of the DAQ in the model represents a novel method of more complete simulations of tomograph designs, and could prove to be of pivotal importance in the optimization of such designs.

  9. [2D imaging simulations of a small animal PET scanner with DOI measurement: jPET-RD.].

    PubMed

    Yamaya, Taiga; Kitamura, Keishi; Hagiwara, Naoki; Obi, Takashi; Hasegawa, Tomoyuki; Yoshida, Eiji; Tsuda, Tomoaki; Inadama, Naoko; Wada, Yasuhiro; Murayama, Hideo

    2005-01-01

    We present a preliminary study on the design of a high sensitivity small animal DOI-PET scanner: jPET-RD (for Rodents with DOI detectors), which will contribute to molecular imaging. The 4-layer DOI block detector for the jPET-RD that consists of scintillation crystals (1.4 mm x 1.4 mm x 4.5 mm) and a flat panel position-sensitive photomultiplier tube (52 mm x 52 mm) was previously proposed. In this paper, we investigate imaging performance of the jPET-RD through numerical simulations. The scanner has a hexagonal geometry with a small diameter and a large axial aperture. Therefore DOI information is expected to improve resolution uniformity in the whole field of view (FOV). We simulate the scanner for various parameters of the number of DOI channels and the crystal length. Simulated data are reconstructed using the maximum likelihood expectation maximization with accurate system modeling. The trade-off results between background noise and spatial resolution show that only shortening the length of crystal does not improve the trade-off at all, and that 4-layer DOI information improves uniformity of spatial resolution in the whole FOV. Excellent performance of the jPET-RD can be expected based on the numerical simulation results. PMID:15961924

  10. Image reconstruction for PET/CT scanners: past achievements and future challenges

    PubMed Central

    Tong, Shan; Alessio, Adam M; Kinahan, Paul E

    2011-01-01

    PET is a medical imaging modality with proven clinical value for disease diagnosis and treatment monitoring. The integration of PET and CT on modern scanners provides a synergy of the two imaging modalities. Through different mathematical algorithms, PET data can be reconstructed into the spatial distribution of the injected radiotracer. With dynamic imaging, kinetic parameters of specific biological processes can also be determined. Numerous efforts have been devoted to the development of PET image reconstruction methods over the last four decades, encompassing analytic and iterative reconstruction methods. This article provides an overview of the commonly used methods. Current challenges in PET image reconstruction include more accurate quantitation, TOF imaging, system modeling, motion correction and dynamic reconstruction. Advances in these aspects could enhance the use of PET/CT imaging in patient care and in clinical research studies of pathophysiology and therapeutic interventions. PMID:21339831

  11. Timing properties measurements of STMicroelectronics silicon photomultipliers for PET scanners

    NASA Astrophysics Data System (ADS)

    Sanfilippo, D.; Fallica, P. G.; Carbone, B.; Mazzillo, M.; Piana, A.; Valvo, G.; La Rocca, P.; Riggi, F.

    2013-02-01

    Preliminary results concerning the study of timing and energy resolution properties of recent Silicon Photomultipliers (SiPMs) devices by STMicroelectronics are here discussed in view of their possible applications in TOF PET modules and PET/MRI hybrid systems. A pair of devices with 60×60 microcells and an active area of 3.5×3.5 mm2 have been coupled to 3×3×15 mm3 LYSO crystals and fully characterized. Measurements of the linearity response with different gamma sources and of the timing resolution with the two 511 keV photons from a 22Na positron source have been carried out.

  12. Small PET scanner based on MRI-compatible light sensor

    NASA Astrophysics Data System (ADS)

    Molnar, J.; Balkay, L.; Berenyi, E.

    2015-03-01

    Improving the quality of life of elderly people requires diagnostic and therapeutic capabilities for diseases of the central nervous system, such as Alzheimer's, Parkinson's, and epilepsy which have a rapidly growing impact on society. Minimallyinvasive imaging technologies such as PET and MRI allow for monitoring and tracking these illnesses, starting from their preliminary manifestations.

  13. Sensitivity booster for DOI-PET scanner by utilizing Compton scattering events between detector blocks

    NASA Astrophysics Data System (ADS)

    Yoshida, Eiji; Tashima, Hideaki; Yamaya, Taiga

    2014-11-01

    In a conventional PET scanner, coincidence events are measured with a limited energy window for detection of photoelectric events in order to reject Compton scatter events that occur in a patient, but Compton scatter events caused in detector crystals are also rejected. Scatter events within the patient causes scatter coincidences, but inter crystal scattering (ICS) events have useful information for determining an activity distribution. Some researchers have reported the feasibility of PET scanners based on a Compton camera for tracing ICS into the detector. However, these scanners require expensive semiconductor detectors for high-energy resolution. In the Anger-type block detector, single photons interacting with multiple detectors can be obtained for each interacting position and complete information can be gotten just as for photoelectric events in the single detector. ICS events in the single detector have been used to get coincidence, but single photons interacting with multiple detectors have not been used to get coincidence. In this work, we evaluated effect of sensitivity improvement using Compton kinetics in several types of DOI-PET scanners. The proposed method promises to improve the sensitivity using coincidence events of single photons interacting with multiple detectors, which are identified as the first interaction (FI). FI estimation accuracy can be improved to determine FI validity from the correlation between Compton scatter angles calculated on the coincidence line-of-response. We simulated an animal PET scanner consisting of 42 detectors. Each detector block consists of three types of scintillator crystals (LSO, GSO and GAGG). After the simulation, coincidence events are added as information for several depth-of-interaction (DOI) resolutions. From the simulation results, we concluded the proposed method promises to improve the sensitivity considerably when effective atomic number of a scintillator is low. Also, we showed that FI estimate

  14. Assessment of the Contrast to Noise Ratio in PET Scanners with Monte Carlo Methods

    NASA Astrophysics Data System (ADS)

    Michail, C. M.; Karpetas, G. E.; Fountos, G. P.; Valais, I. G.; Nikolopoulos, D.; Kandarakis, I. S.; Panayiotakis, G. S.

    2015-09-01

    The aim of the present study was to assess the contrast to noise ratio (CNR) of PET scanners through a thin layer chromatography (TLC) plane source. The source was simulated using a previously validated Monte Carlo model. The model was developed by using the GATE MC package and reconstructed images obtained with the STIR software for tomographic image reconstruction. The PET scanner simulated was the GE DiscoveryST. A plane source consisted of a TLC plate, was simulated by a layer of silica gel on aluminum (Al) foil substrates, immersed in 18F-FDG bath solution. Image quality was assessed in terms of the CNR. CNR was estimated from coronal reconstructed images of the plane source. Images were reconstructed by the maximum likelihood estimation (MLE)-OSMAPOSL. OSMAPOSL reconstruction was assessed by using various subsets (3, 15 and 21) and various iterations (2 to 20). CNR values were found to decrease when both iterations and subsets increase. Two (2) iterations were found to be optimal. The simulated PET evaluation method, based on the TLC plane source, can be useful in image quality assessment of PET scanners.

  15. Waveform-Sampling Electronics for a Whole-Body Time-of-Flight PET Scanner

    PubMed Central

    Ashmanskas, W. J.; LeGeyt, B. C.; Newcomer, F. M.; Panetta, J. V.; Ryan, W. A.; Van Berg, R.; Wiener, R. I.; Karp Fellow, J. S.

    2014-01-01

    Waveform sampling is an appealing technique for instruments requiring precision time and pulse-height measurements. Sampling each PMT waveform at oscilloscope-like rates of several gigasamples per second enables one to process PMT signals digitally, which in turn makes it straightforward to optimize timing resolution and amplitude (energy and position) resolution in response to calibration effects, pile-up effects, and other systematic sources of waveform variation. We describe a system design and preliminary implementation that neatly maps waveform-sampling technology onto the LaPET prototype whole-body time-of-flight PET scanner that serves as the platform for testing this new technology. PMID:25484379

  16. Using triple gamma coincidences with a pixelated semiconductor Compton-PET scanner: a simulation study

    NASA Astrophysics Data System (ADS)

    Kolstein, M.; Chmeissani, M.

    2016-01-01

    The Voxel Imaging PET (VIP) Pathfinder project presents a novel design using pixelated semiconductor detectors for nuclear medicine applications to achieve the intrinsic image quality limits set by physics. The conceptual design can be extended to a Compton gamma camera. The use of a pixelated CdTe detector with voxel sizes of 1 × 1 × 2 mm3 guarantees optimal energy and spatial resolution. However, the limited time resolution of semiconductor detectors makes it impossible to use Time Of Flight (TOF) with VIP PET. TOF is used in order to improve the signal to noise ratio (SNR) by using only the most probable portion of the Line-Of-Response (LOR) instead of its entire length. To overcome the limitation of CdTe time resolution, we present in this article a simulation study using β+-γ emitting isotopes with a Compton-PET scanner. When the β+ annihilates with an electron it produces two gammas which produce a LOR in the PET scanner, while the additional gamma, when scattered in the scatter detector, provides a Compton cone that intersects with the aforementioned LOR. The intersection indicates, within a few mm of uncertainty along the LOR, the origin of the beta-gamma decay. Hence, one can limit the part of the LOR used by the image reconstruction algorithm.

  17. Raytest ClearPET™, a new generation small animal PET scanner

    NASA Astrophysics Data System (ADS)

    Sempere Roldan, P.; Chereul, E.; Dietzel, O.; Magnier, L.; Pautrot, C.; Rbah, L.; Sappey-Marinier, D.; Wagner, A.; Zimmer, L.; Janier, M.; Tarazona, V.; Dietzel, G.

    2007-02-01

    The ClearPET is a high-performance small animal PET scanner that has been developed by the Crystal Clear Collaboration (CCC) and is now commercially available to customers worldwide through the Raytest group. Its high sensitivity and spatial resolution are achieved, thanks to a patented phoswich configuration made of two types of scintillating crystals: L(Y)SO and LuYAP:Ce, thus providing depth of interaction information. The ClearPET presents a modular design and it allows for 360° rotation of the detector modules around the field of view (FOV). The ClearPET also presents the unique feature of having an adjustable FOV diameter, thus being suitable for small and medium size animal PET studies. This paper shows the performances of the commercial ClearPET in terms of its spatial resolution, sensitivity and quality of phantoms and preclinical images obtained. It also describes the updated situation of the first commercial unit sold to the Animage platform in Lyon.

  18. Pragmatic fully 3D image reconstruction for the MiCES mouse imaging PET scanner

    NASA Astrophysics Data System (ADS)

    Lee, Kisung; Kinahan, Paul E.; Fessler, Jeffrey A.; Miyaoka, Robert S.; Janes, Marie; Lewellen, Tom K.

    2004-10-01

    We present a pragmatic approach to image reconstruction for data from the micro crystal elements system (MiCES) fully 3D mouse imaging positron emission tomography (PET) scanner under construction at the University of Washington. Our approach is modelled on fully 3D image reconstruction used in clinical PET scanners, which is based on Fourier rebinning (FORE) followed by 2D iterative image reconstruction using ordered-subsets expectation-maximization (OSEM). The use of iterative methods allows modelling of physical effects (e.g., statistical noise, detector blurring, attenuation, etc), while FORE accelerates the reconstruction process by reducing the fully 3D data to a stacked set of independent 2D sinograms. Previous investigations have indicated that non-stationary detector point-spread response effects, which are typically ignored for clinical imaging, significantly impact image quality for the MiCES scanner geometry. To model the effect of non-stationary detector blurring (DB) in the FORE+OSEM(DB) algorithm, we have added a factorized system matrix to the ASPIRE reconstruction library. Initial results indicate that the proposed approach produces an improvement in resolution without an undue increase in noise and without a significant increase in the computational burden. The impact on task performance, however, remains to be evaluated.

  19. Quantitative comparison of FBP, EM, and Bayesian reconstruction algorithms for the IndyPET scanner.

    PubMed

    Frese, Thomas; Rouze, Ned C; Bouman, Charles A; Sauer, Ken; Hutchins, Gary D

    2003-02-01

    We quantitatively compare filtered backprojection (FBP), expectation-maximization (EM), and Bayesian reconstruction algorithms as applied to the IndyPET scanner--a dedicated research scanner which has been developed for small and intermediate field of view imaging applications. In contrast to previous approaches that rely on Monte Carlo simulations, a key feature of our investigation is the use of an empirical system kernel determined from scans of line source phantoms. This kernel is incorporated into the forward model of the EM and Bayesian algorithms to achieve resolution recovery. Three data sets are used, data collected on the IndyPET scanner using a bar phantom and a Hoffman three-dimensional brain phantom, and simulated data containing a hot lesion added to a uniform background. Reconstruction quality is analyzed quantitatively in terms of bias-variance measures (bar phantom) and mean square error (lesion phantom). We observe that without use of the empirical system kernel, the FBP, EM, and Bayesian algorithms give similar performance. However, with the inclusion of the empirical kernel, the iterative algorithms provide superior reconstructions compared with FBP, both in terms of visual quality and quantitative measures. Furthermore, Bayesian methods outperform EM. We conclude that significant improvements in reconstruction quality can be realized by combining accurate models of the system response with Bayesian reconstruction algorithms. PMID:12716002

  20. Crystal timing offset calibration method for time of flight PET scanners

    NASA Astrophysics Data System (ADS)

    Ye, Jinghan; Song, Xiyun

    2016-03-01

    In time-of-flight (TOF) positron emission tomography (PET), precise calibration of the timing offset of each crystal of a PET scanner is essential. Conventionally this calibration requires a specially designed tool just for this purpose. In this study a method that uses a planar source to measure the crystal timing offsets (CTO) is developed. The method uses list mode acquisitions of a planar source placed at multiple orientations inside the PET scanner field-of-view (FOV). The placement of the planar source in each acquisition is automatically figured out from the measured data, so that a fixture for exactly placing the source is not required. The expected coincidence time difference for each detected list mode event can be found from the planar source placement and the detector geometry. A deviation of the measured time difference from the expected one is due to CTO of the two crystals. The least squared solution of the CTO is found iteratively using the list mode events. The effectiveness of the crystal timing calibration method is evidenced using phantom images generated by placing back each list mode event into the image space with the timing offset applied to each event. The zigzagged outlines of the phantoms in the images become smooth after the crystal timing calibration is applied. In conclusion, a crystal timing calibration method is developed. The method uses multiple list mode acquisitions of a planar source to find the least squared solution of crystal timing offsets.

  1. Monte Carlo modelling of singles-mode transmission data for small animal PET scanners

    NASA Astrophysics Data System (ADS)

    Vandervoort, Eric; Camborde, Marie-Laure; Jan, Sébastien; Sossi, Vesna

    2007-06-01

    The attenuation corrections factors (ACFs), which are necessary for quantitatively accurate PET imaging, can be obtained using singles-mode transmission scanning. However, contamination from scatter is a largely unresolved problem for these data. We present an extension of the Monte Carlo simulation tool, GATE, for singles-mode transmission data and its validation using experimental data from the microPET R4 and Focus 120 scanners. We first validated our simulated PET scanner for coincidence-mode data where we found that experimental resolution and scatter fractions (SFs) agreed well for simulations that included positron interactions and scatter in the source material. After modifying GATE to model singles-mode data, we compared simulated and experimental ACFs and SFs for three different sized water cylinders using 57Co (122 keV photon emitter) and 68Ge (positron emitter) transmission sources. We also propose a simple correction for a large background contamination we identified in the 68Ge singles-mode data due to intrinsic 176Lu radioactivity present in the detector crystals. For simulation data, the SFs agreed to within 1.5% and 2.5% of experimental values for background-corrected 68Ge and 57Co transmission data, respectively. This new simulation tool accurately models the photon interactions and data acquisition for singles-mode transmission scans.

  2. Errors in MR-based attenuation correction for brain imaging with PET/MR scanners

    NASA Astrophysics Data System (ADS)

    Rota Kops, Elena; Herzog, Hans

    2013-02-01

    AimAttenuation correction of PET data acquired by hybrid MR/PET scanners remains a challenge, even if several methods for brain and whole-body measurements have been developed recently. A template-based attenuation correction for brain imaging proposed by our group is easy to handle and delivers reliable attenuation maps in a short time. However, some potential error sources are analyzed in this study. We investigated the choice of template reference head among all the available data (error A), and possible skull anomalies of the specific patient, such as discontinuities due to surgery (error B). Materials and methodsAn anatomical MR measurement and a 2-bed-position transmission scan covering the whole head and neck region were performed in eight normal subjects (4 females, 4 males). Error A: Taking alternatively one of the eight heads as reference, eight different templates were created by nonlinearly registering the images to the reference and calculating the average. Eight patients (4 females, 4 males; 4 with brain lesions, 4 w/o brain lesions) were measured in the Siemens BrainPET/MR scanner. The eight templates were used to generate the patients' attenuation maps required for reconstruction. ROI and VOI atlas-based comparisons were performed employing all the reconstructed images. Error B: CT-based attenuation maps of two volunteers were manipulated by manually inserting several skull lesions and filling a nasal cavity. The corresponding attenuation coefficients were substituted with the water's coefficient (0.096/cm). ResultsError A: The mean SUVs over the eight templates pairs for all eight patients and all VOIs did not differ significantly one from each other. Standard deviations up to 1.24% were found. Error B: After reconstruction of the volunteers' BrainPET data with the CT-based attenuation maps without and with skull anomalies, a VOI-atlas analysis was performed revealing very little influence of the skull lesions (less than 3%), while the filled nasal

  3. Comparative evaluation of two commercial PET scanners, ECAT EXACT HR+ and Biograph 2, using GATE

    NASA Astrophysics Data System (ADS)

    Karakatsanis, N.; Sakellios, N.; Tsantilas, N. X.; Dikaios, N.; Tsoumpas, C.; Lazaro, D.; Loudos, G.; Schmidtlein, C. R.; Louizi, K.; Valais, J.; Nikolopoulos, D.; Malamitsi, J.; Kandarakis, J.; Nikita, K.

    2006-12-01

    Geant4 application for tomographic emission (GATE) is a generic Monte Carlo simulation platform based on a general-purpose code GEANT4 and designed to simulate positron emission tomography (PET) and single photon emission tomography systems. Monte Carlo simulations are used in nuclear medicine to model imaging systems and develop and assess tomographic reconstruction algorithms and correction methods for improved image quantification. The purpose of this study is to validate two GATE models of the commercial available PET scanner HR+ and the PET/CT Biograph 2. The geometry of the system components has been described in GATE, including detector ring, crystal blocks, PMTs etc. The energy and spatial resolution of the scanners as given by the manufacturers have been taken into account. The GATE simulated results are compared directly to experimental data obtained using a number of NEMA NU-2-2001 performance protocols, including spatial resolution, sensitivity and scatter fraction. All the respective phantoms are precisely modeled. Furthermore, an approximate dead-time model both at the level of single and coincidence events was developed so that the simulated count rate curve can satisfactorily match the experimental count rate performance curve for each scanner In addition a software tool was developed to build the sinograms from the simulated data and import them into the software for tomographic image reconstruction where the reconstruction algorithm of FBP3DRP was applied. An agreement of less than 0.8 mm was obtained between the spatial resolution of the simulated system and the experimental results. Also the simulated scatter fraction for the NEMA NU 2-2001 scatter phantom matched the experimental results to within 3% of measured values. Finally the ratio of the simulated sensitivities with sources radially offset 0 and 10 cm from the central axis of each of the two scanners reaches an agreement of less than 1% between the simulated and experimental values. This

  4. Preliminary assessment of the imaging capability of the YAP (S)PET small animal scanner in neuroscience

    NASA Astrophysics Data System (ADS)

    Bartoli, Antonietta; Belcari, Nicola; Stark, Daniela; Höhnemann, Sabine; Piel, Markus; Jennewein, Marc; Schmitt, Ulrich; Tillmanns, Julia; Thews, Oliver; Hiemke, Christoph; Roesch, Frank; Del Guerra, Alberto

    2006-12-01

    The new and fully engineered version of the YAP-(S)PET small animal scanner has been tested at the University of Mainz for preliminary assessment of its imaging capability for studies related to neuropharmacology and psychiatry. The main feature of the scanner is the capability to combine PET and SPECT techniques. It allows the development of new and interesting protocols for the investigation of many biological phenomena, more effectively than with PET or SPECT modalities alone. The scanner is made up of four detector heads, each one composed of a 4×4 cm 2 of YAlO 3:Ce (or YAP:Ce) matrix, and has a field of view (FOV) of 4 cm axially×4 cm ∅ transaxially. In PET mode, the volume resolution is less than 8 mm 3 and is nearly constant over the whole FOV, while the sensitivity is about 2%. The SPECT performance is not so good, due to the presence of the multi-hole lead collimator in front of each head. Nevertheless, the YAP-PET scanner offers excellent resolution and sensitivity for performing on the availability of D2-like dopamine receptors on mice and rats in both PET and SPECT modalities.

  5. Effect of filters and reconstruction algorithms on I-124 PET in Siemens Inveon PET scanner

    NASA Astrophysics Data System (ADS)

    Ram Yu, A.; Kim, Jin Su

    2015-10-01

    Purpose: To assess the effects of filtering and reconstruction on Siemens I-124 PET data. Methods: A Siemens Inveon PET was used. Spatial resolution of I-124 was measured to a transverse offset of 50 mm from the center FBP, 2D ordered subset expectation maximization (OSEM2D), 3D re-projection algorithm (3DRP), and maximum a posteriori (MAP) methods were tested. Non-uniformity (NU), recovery coefficient (RC), and spillover ratio (SOR) parameterized image quality. Mini deluxe phantom data of I-124 was also assessed. Results: Volumetric resolution was 7.3 mm3 from the transverse FOV center when FBP reconstruction algorithms with ramp filter was used. MAP yielded minimal NU with β =1.5. OSEM2D yielded maximal RC. SOR was below 4% for FBP with ramp, Hamming, Hanning, or Shepp-Logan filters. Based on the mini deluxe phantom results, an FBP with Hanning or Parzen filters, or a 3DRP with Hanning filter yielded feasible I-124 PET data.Conclusions: Reconstruction algorithms and filters were compared. FBP with Hanning or Parzen filters, or 3DRP with Hanning filter yielded feasible data for quantifying I-124 PET.

  6. Using compressive sensing to recover images from PET scanners with partial detector rings

    SciTech Connect

    Valiollahzadeh, SeyyedMajid; Clark, John W.; Mawlawi, Osama

    2015-01-15

    Purpose: Most positron emission tomography/computed tomography (PET/CT) scanners consist of tightly packed discrete detector rings to improve scanner efficiency. The authors’ aim was to use compressive sensing (CS) techniques in PET imaging to investigate the possibility of decreasing the number of detector elements per ring (introducing gaps) while maintaining image quality. Methods: A CS model based on a combination of gradient magnitude and wavelet domains (wavelet-TV) was developed to recover missing observations in PET data acquisition. The model was designed to minimize the total variation (TV) and L1-norm of wavelet coefficients while constrained by the partially observed data. The CS model also incorporated a Poisson noise term that modeled the observed noise while suppressing its contribution by penalizing the Poisson log likelihood function. Three experiments were performed to evaluate the proposed CS recovery algorithm: a simulation study, a phantom study, and six patient studies. The simulation dataset comprised six disks of various sizes in a uniform background with an activity concentration of 5:1. The simulated image was multiplied by the system matrix to obtain the corresponding sinogram and then Poisson noise was added. The resultant sinogram was masked to create the effect of partial detector removal and then the proposed CS algorithm was applied to recover the missing PET data. In addition, different levels of noise were simulated to assess the performance of the proposed algorithm. For the phantom study, an IEC phantom with six internal spheres each filled with F-18 at an activity-to-background ratio of 10:1 was used. The phantom was imaged twice on a RX PET/CT scanner: once with all detectors operational (baseline) and once with four detector blocks (11%) turned off at each of 0 °, 90 °, 180 °, and 270° (partially sampled). The partially acquired sinograms were then recovered using the proposed algorithm. For the third test, PET images

  7. NEMA and clinical evaluation of a novel brain PET-CT scanner

    PubMed Central

    Grogg, Kira S.; Toole, Terrence; Ouyang, Jinsong; Zhu, Xuping; Normandin, Marc; Johnson, Keith; Alpert, Nathaniel M.; Fakhri, Georges El

    2016-01-01

    The aim of this study was to determine the performance of a novel mobile human brain/small animal PET-CT system, developed by Photo Diagnostic Systems Inc. The scanner has a 35.7-cm diameter bore and a 22-cm axial extent. The detector ring has 7 modules each with 3×4 cerium-doped lutetium yttrium orthosilicate crystal blocks, each consisting of 22×22 outer layer and 21×21 inner layer crystals, each layer 1 cm thick. Light is collected by 12×12 SiPMs. The integrated CT can be used for attenuation correction and anatomical localization. The scanner was designed as a low-cost device that nevertheless produces high-quality PET images with the unique capability of battery-powered propulsion, enabling use in many settings. Methods Spatial resolution, sensitivity and noise-equivalent count rate (NECR) were measured based on the National Electrical Manufacturers Association NU2-2012 procedures. Reconstruction was done with tight energy and timing cuts: 400-650 keV and 7ns, and loose cuts: 350-700 keV and 10ns. Additional image quality measurements were made from phantoms, human, and animal studies. Performance was compared to a reference scanner (ECAT Exact HR+) with comparable imaging properties. Results The full-width half-max transverse resolution at 1 cm (10 cm) radius is 3.2 mm (5.2 mm radial, 3.1 mm tangential) and the axial resolution is 3.5 mm (4.0 mm). For tight (loose) cuts, a sensitivity of 7.5 (11.7) kcps/MBq at the center increases to 8.8 (13.9) kcps/MBq at a 10 cm radial offset. The maximum NECR of 19.5 (22.7) kcps was achieved for an activity concentration of 2.9 kBq/ml. Contrast recovery for 4:1 hot cylinder to warm background was 76% for the 25 mm diameter cylinder, but decreased with decreasing cylinder size. The quantitation agrees within 2% of the known activity distribution and concentration. Brain phantom and human scans have shown agreement in SUV values and image quality with the HR+. Conclusion We have characterized the performance of the NeuroPET

  8. Design of a Second Generation Firewire Based Data Acquisition System for Small Animal PET Scanners

    PubMed Central

    Lewellen, T.K.; Miyaoka, R.S.; MacDonald, L.R.; Haselman, M.; DeWitt, D.; Hunter, William; Hauck, S.

    2009-01-01

    The University of Washington developed a Firewire based data acquisition system for the MiCES small animal PET scanner. Development work has continued on new imaging scanners that require more data channels and need to be able to operate within a MRI imaging system. To support these scanners, we have designed a new version of our data acquisition system that leverages the capabilities of modern field programmable gate arrays (FPGA). The new design preserves the basic approach of the original system, but puts almost all functions into the FPGA, including the Firewire elements, the embedded processor, and pulse timing and pulse integration. The design has been extended to support implementation of the position estimation and DOl algorithms developed for the cMiCE detector module. The design is centered around an acquisition node board (ANB) that includes 65 ADC channels, Firewire 1394b support, the FPGA, a serial command bus and signal lines to support a rough coincidence window implementation to reject singles events from being sent on the Firewire bus. Adapter boards convert detector signals into differential paired signals to connect to the ANB. PMID:20228958

  9. A feasibility study of ortho-positronium decays measurement with the J-PET scanner based on plastic scintillators

    NASA Astrophysics Data System (ADS)

    Kamińska, D.; Gajos, A.; Czerwiński, E.; Alfs, D.; Bednarski, T.; Białas, P.; Curceanu, C.; Dulski, K.; Głowacz, B.; Gupta-Sharma, N.; Gorgol, M.; Hiesmayr, B. C.; Jasińska, B.; Korcyl, G.; Kowalski, P.; Krzemień, W.; Krawczyk, N.; Kubicz, E.; Mohammed, M.; Niedźwiecki, Sz.; Pawlik-Niedźwiecka, M.; Raczyński, L.; Rudy, Z.; Silarski, M.; Wieczorek, A.; Wiślicki, W.; Zgardzińska, B.; Zieliński, M.; Moskal, P.

    2016-08-01

    We present a study of the application of the Jagiellonian positron emission tomograph (J-PET) for the registration of gamma quanta from decays of ortho-positronium (o-Ps). The J-PET is the first positron emission tomography scanner based on organic scintillators in contrast to all current PET scanners based on inorganic crystals. Monte Carlo simulations show that the J-PET as an axially symmetric and high acceptance scanner can be used as a multi-purpose detector well suited to pursue research including e.g. tests of discrete symmetries in decays of ortho-positronium in addition to the medical imaging. The gamma quanta originating from o-Ps decay interact in the plastic scintillators predominantly via the Compton effect, making the direct measurement of their energy impossible. Nevertheless, it is shown in this paper that the J-PET scanner will enable studies of the { o-Ps }→ 3γ decays with angular and energy resolution equal to σ (θ ) ≈ {0.4°} and σ (E) ≈ 4.1 {keV}, respectively. An order of magnitude shorter decay time of signals from plastic scintillators with respect to the inorganic crystals results not only in better timing properties crucial for the reduction of physical and instrumental background, but also suppresses significantly the pile-ups, thus enabling compensation of the lower efficiency of the plastic scintillators by performing measurements with higher positron source activities.

  10. Designing a compact high performance brain PET scanner-simulation study.

    PubMed

    Gong, Kuang; Majewski, Stan; Kinahan, Paul E; Harrison, Robert L; Elston, Brian F; Manjeshwar, Ravindra; Dolinsky, Sergei; Stolin, Alexander V; Brefczynski-Lewis, Julie A; Qi, Jinyi

    2016-05-21

    The desire to understand normal and disordered human brain function of upright, moving persons in natural environments motivates the development of the ambulatory micro-dose brain PET imager (AMPET). An ideal system would be light weight but with high sensitivity and spatial resolution, although these requirements are often in conflict with each other. One potential approach to meet the design goals is a compact brain-only imaging device with a head-sized aperture. However, a compact geometry increases parallax error in peripheral lines of response, which increases bias and variance in region of interest (ROI) quantification. Therefore, we performed simulation studies to search for the optimal system configuration and to evaluate the potential improvement in quantification performance over existing scanners. We used the Cramér-Rao variance bound to compare the performance for ROI quantification using different scanner geometries. The results show that while a smaller ring diameter can increase photon detection sensitivity and hence reduce the variance at the center of the field of view, it can also result in higher variance in peripheral regions when the length of detector crystal is 15 mm or more. This variance can be substantially reduced by adding depth-of-interaction (DOI) measurement capability to the detector modules. Our simulation study also shows that the relative performance depends on the size of the ROI, and a large ROI favors a compact geometry even without DOI information. Based on these results, we propose a compact 'helmet' design using detectors with DOI capability. Monte Carlo simulations show the helmet design can achieve four-fold higher sensitivity and resolve smaller features than existing cylindrical brain PET scanners. The simulations also suggest that improving TOF timing resolution from 400 ps to 200 ps also results in noticeable improvement in image quality, indicating better timing resolution is desirable for brain imaging. PMID

  11. Time-invariant component-based normalization for a simultaneous PET-MR scanner.

    PubMed

    Belzunce, M A; Reader, A J

    2016-05-01

    Component-based normalization is a method used to compensate for the sensitivity of each of the lines of response acquired in positron emission tomography. This method consists of modelling the sensitivity of each line of response as a product of multiple factors, which can be classified as time-invariant, time-variant and acquisition-dependent components. Typical time-variant factors are the intrinsic crystal efficiencies, which are needed to be updated by a regular normalization scan. Failure to do so would in principle generate artifacts in the reconstructed images due to the use of out of date time-variant factors. For this reason, an assessment of the variability and the impact of the crystal efficiencies in the reconstructed images is important to determine the frequency needed for the normalization scans, as well as to estimate the error obtained when an inappropriate normalization is used. Furthermore, if the fluctuations of these components are low enough, they could be neglected and nearly artifact-free reconstructions become achievable without performing a regular normalization scan. In this work, we analyse the impact of the time-variant factors in the component-based normalization used in the Biograph mMR scanner, but the work is applicable to other PET scanners. These factors are the intrinsic crystal efficiencies and the axial factors. For the latter, we propose a new method to obtain fixed axial factors that was validated with simulated data. Regarding the crystal efficiencies, we assessed their fluctuations during a period of 230 d and we found that they had good stability and low dispersion. We studied the impact of not including the intrinsic crystal efficiencies in the normalization when reconstructing simulated and real data. Based on this assessment and using the fixed axial factors, we propose the use of a time-invariant normalization that is able to achieve comparable results to the standard, daily updated, normalization factors used in this

  12. Time-invariant component-based normalization for a simultaneous PET-MR scanner

    NASA Astrophysics Data System (ADS)

    Belzunce, M. A.; Reader, A. J.

    2016-05-01

    Component-based normalization is a method used to compensate for the sensitivity of each of the lines of response acquired in positron emission tomography. This method consists of modelling the sensitivity of each line of response as a product of multiple factors, which can be classified as time-invariant, time-variant and acquisition-dependent components. Typical time-variant factors are the intrinsic crystal efficiencies, which are needed to be updated by a regular normalization scan. Failure to do so would in principle generate artifacts in the reconstructed images due to the use of out of date time-variant factors. For this reason, an assessment of the variability and the impact of the crystal efficiencies in the reconstructed images is important to determine the frequency needed for the normalization scans, as well as to estimate the error obtained when an inappropriate normalization is used. Furthermore, if the fluctuations of these components are low enough, they could be neglected and nearly artifact-free reconstructions become achievable without performing a regular normalization scan. In this work, we analyse the impact of the time-variant factors in the component-based normalization used in the Biograph mMR scanner, but the work is applicable to other PET scanners. These factors are the intrinsic crystal efficiencies and the axial factors. For the latter, we propose a new method to obtain fixed axial factors that was validated with simulated data. Regarding the crystal efficiencies, we assessed their fluctuations during a period of 230 d and we found that they had good stability and low dispersion. We studied the impact of not including the intrinsic crystal efficiencies in the normalization when reconstructing simulated and real data. Based on this assessment and using the fixed axial factors, we propose the use of a time-invariant normalization that is able to achieve comparable results to the standard, daily updated, normalization factors used in this

  13. Detector development for microPET II: a 1 microl resolution PET scanner for small animal imaging.

    PubMed

    Chatziioannou, A; Tai, Y C; Doshi, N; Cherry, S R

    2001-11-01

    We are currently developing a small animal positron emission tomography (PET) scanner with a design goal of 1 microlitre (1 mm3) image resolution. The detectors consist of a 12 x 12 array of 1 x 1 x 10 mm lutetium oxyorthosilicate (LSO) scintillator crystals coupled to a 64-channel photomultiplier tube (PMT) via 5 cm long optical fibre bundles. The optical fibre connection allows a high detector packing fraction despite the dead space surrounding the active region of the PMT. Optical fibre bundles made from different types of glass were tested for light transmission, and also their effects on crystal identification and energy resolution, and compared to direct coupling of the LSO arrays to the PMTs. We also investigated the effects of extramural absorber (EMA) in the fibre bundles. Based on these results, fibre bundles manufactured from F2 glass were selected. We built three pairs of prototype detectors (directly coupled LSO array, fibre bundle without EMA and fibre bundle with EMA) and measured flood histograms, energy resolution, intrinsic spatial resolution and timing resolution. The results demonstrated an intrinsic spatial resolution (FWHM) of 1.12 mm (directly coupled), 1.23 mm (fibre bundle without EMA coupling) and 1.27 mm (fibre bundle with EMA coupling) using an approximately 500 microm diameter Na-22 point source. Using a 330 microm outer diameter steel needle line source filled with F-18, spatial resolution for the detector with the EMA optical fibre bundle improved to 1.05 mm. The respective timing and energy FWHM values were 1.96 ns, 21% (directly coupled), 2.20 ns, 23% (fibre bundle without EMA) and 2.99 ns, 30% (fibre bundle with EMA). The peak-to-valley ratio in the flood histograms was better with EMA (5:1) compared to the optical fibre bundle without EMA (2.5:1), due to the decreased optical cross-talk. In comparison to the detectors used in our current generation microPET scanner, these detectors substantially improve on the spatial resolution

  14. Performance evaluation and optimization of the MiniPET-II scanner

    NASA Astrophysics Data System (ADS)

    Lajtos, Imre; Emri, Miklos; Kis, Sandor A.; Opposits, Gabor; Potari, Norbert; Kiraly, Beata; Nagy, Ferenc; Tron, Lajos; Balkay, Laszlo

    2013-04-01

    This paper presents results of the performance of a small animal PET system (MiniPET-II) installed at our Institute. MiniPET-II is a full ring camera that includes 12 detector modules in a single ring comprised of 1.27×1.27×12 mm3 LYSO scintillator crystals. The axial field of view and the inner ring diameter are 48 mm and 211 mm, respectively. The goal of this study was to determine the NEMA-NU4 performance parameters of the scanner. In addition, we also investigated how the calculated parameters depend on the coincidence time window (τ=2, 3 and 4 ns) and the low threshold settings of the energy window (Elt=250, 350 and 450 keV). Independent measurements supported optimization of the effective system radius and the coincidence time window of the system. We found that the optimal coincidence time window and low threshold energy window are 3 ns and 350 keV, respectively. The spatial resolution was close to 1.2 mm in the center of the FOV with an increase of 17% at the radial edge. The maximum value of the absolute sensitivity was 1.37% for a point source. Count rate tests resulted in peak values for the noise equivalent count rate (NEC) curve and scatter fraction of 14.2 kcps (at 36 MBq) and 27.7%, respectively, using the rat phantom. Numerical values of the same parameters obtained for the mouse phantom were 55.1 kcps (at 38.8 MBq) and 12.3%, respectively. The recovery coefficients of the image quality phantom ranged from 0.1 to 0.87. Altering the τ and Elt resulted in substantial changes in the NEC peak and the sensitivity while the effect on the image quality was negligible. The spatial resolution proved to be, as expected, independent of the τ and Elt. The calculated optimal effective system radius (resulting in the best image quality) was 109 mm. Although the NEC peak parameters do not compare favorably with those of other small animal scanners, it can be concluded that under normal counting situations the MiniPET-II imaging capability assures remarkably

  15. NEMA NU 4-2008 validation and applications of the PET-SORTEO Monte Carlo simulations platform for the geometry of the Inveon PET preclinical scanner.

    PubMed

    Boisson, F; Wimberley, C J; Lehnert, W; Zahra, D; Pham, T; Perkins, G; Hamze, H; Gregoire, M-C; Reilhac, A

    2013-10-01

    Monte Carlo-based simulation of positron emission tomography (PET) data plays a key role in the design and optimization of data correction and processing methods. Our first aim was to adapt and configure the PET-SORTEO Monte Carlo simulation program for the geometry of the widely distributed Inveon PET preclinical scanner manufactured by Siemens Preclinical Solutions. The validation was carried out against actual measurements performed on the Inveon PET scanner at the Australian Nuclear Science and Technology Organisation in Australia and at the Brain & Mind Research Institute and by strictly following the NEMA NU 4-2008 standard. The comparison of simulated and experimental performance measurements included spatial resolution, sensitivity, scatter fraction and count rates, image quality and Derenzo phantom studies. Results showed that PET-SORTEO reliably reproduces the performances of this Inveon preclinical system. In addition, imaging studies showed that the PET-SORTEO simulation program provides raw data for the Inveon scanner that can be fully corrected and reconstructed using the same programs as for the actual data. All correction techniques (attenuation, scatter, randoms, dead-time, and normalization) can be applied on the simulated data leading to fully quantitative reconstructed images. In the second part of the study, we demonstrated its ability to generate fast and realistic biological studies. PET-SORTEO is a workable and reliable tool that can be used, in a classical way, to validate and/or optimize a single PET data processing step such as a reconstruction method. However, we demonstrated that by combining a realistic simulated biological study ([(11)C]Raclopride here) involving different condition groups, simulation allows one also to assess and optimize the data correction, reconstruction and data processing line flow as a whole, specifically for each biological study, which is our ultimate intent. PMID:24018840

  16. NEMA NU 4-2008 validation and applications of the PET-SORTEO Monte Carlo simulations platform for the geometry of the Inveon PET preclinical scanner

    NASA Astrophysics Data System (ADS)

    Boisson, F.; Wimberley, C. J.; Lehnert, W.; Zahra, D.; Pham, T.; Perkins, G.; Hamze, H.; Gregoire, M.-C.; Reilhac, A.

    2013-10-01

    Monte Carlo-based simulation of positron emission tomography (PET) data plays a key role in the design and optimization of data correction and processing methods. Our first aim was to adapt and configure the PET-SORTEO Monte Carlo simulation program for the geometry of the widely distributed Inveon PET preclinical scanner manufactured by Siemens Preclinical Solutions. The validation was carried out against actual measurements performed on the Inveon PET scanner at the Australian Nuclear Science and Technology Organisation in Australia and at the Brain & Mind Research Institute and by strictly following the NEMA NU 4-2008 standard. The comparison of simulated and experimental performance measurements included spatial resolution, sensitivity, scatter fraction and count rates, image quality and Derenzo phantom studies. Results showed that PET-SORTEO reliably reproduces the performances of this Inveon preclinical system. In addition, imaging studies showed that the PET-SORTEO simulation program provides raw data for the Inveon scanner that can be fully corrected and reconstructed using the same programs as for the actual data. All correction techniques (attenuation, scatter, randoms, dead-time, and normalization) can be applied on the simulated data leading to fully quantitative reconstructed images. In the second part of the study, we demonstrated its ability to generate fast and realistic biological studies. PET-SORTEO is a workable and reliable tool that can be used, in a classical way, to validate and/or optimize a single PET data processing step such as a reconstruction method. However, we demonstrated that by combining a realistic simulated biological study ([11C]Raclopride here) involving different condition groups, simulation allows one also to assess and optimize the data correction, reconstruction and data processing line flow as a whole, specifically for each biological study, which is our ultimate intent.

  17. A feasibility study of PETiPIX: an ultra high resolution small animal PET scanner

    NASA Astrophysics Data System (ADS)

    Li, K.; Safavi-Naeini, M.; Franklin, D. R.; Petasecca, M.; Guatelli, S.; Rosenfeld, A. B.; Hutton, B. F.; Lerch, M. L. F.

    2013-12-01

    PETiPIX is an ultra high spatial resolution positron emission tomography (PET) scanner designed for imaging mice brains. Four Timepix pixellated silicon detector modules are placed in an edge-on configuration to form a scanner with a field of view (FoV) 15 mm in diameter. Each detector module consists of 256 × 256 pixels with dimensions of 55 × 55 × 300 μm3. Monte Carlo simulations using GEANT4 Application for Tomographic Emission (GATE) were performed to evaluate the feasibility of the PETiPIX design, including estimation of system sensitivity, angular dependence, spatial resolution (point source, hot and cold phantom studies) and evaluation of potential detector shield designs. Initial experimental work also established that scattered photons and recoil electrons could be detected using a single edge-on Timepix detector with a positron source. Simulation results estimate a spatial resolution of 0.26 mm full width at half maximum (FWHM) at the centre of FoV and 0.29 mm FWHM overall spatial resolution with sensitivity of 0.01%, and indicate that a 1.5 mm thick tungsten shield parallel to the detectors will absorb the majority of non-coplanar annihilation photons, significantly reducing the rates of randoms. Results from the simulated phantom studies demonstrate that PETiPIX is a promising design for studies demanding high resolution images of mice brains.

  18. A high resolution animal PET scanner using compact PS-PMT detectors

    SciTech Connect

    Watanabe, M.; Okada, H.; Shimizu, K.; Omura, T.

    1996-12-31

    A new high resolution PET scanner dedicated to animal studies has been designed, built and tested. The system utilizes 240 block detectors, each of which consists of a new compact position-sensitive photomultiplier tube (PS-PMT) and an 8 x 4 BGO array. A total number of 7,680 crystals (480 per ring) are positioned to form a 508 mm diameter of 16 detector rings with 7.2 mm pitch and 114 mm axial field of view (FOV). The system is designed to perform activation studies using a monkey in a sitting position. The data can be acquired in either 2D or 3D mode, where the slice collimators are retracted in 3D mode. The transaxial resolution is 2.6 mm FWHM at the center of the FOV, and the average axial resolution on the axis of the ring is 3.3 mm FWHM in the direct slice and 3.2 mm FWHM in the cross slice. The scatter fraction, sensitivity and count rate performance were evaluated for a 10 cm diameter cylindrical phantom. The total system sensitivity is 2.3 kcps/kBq/ml in 2D mode and 22.8 kcps/kBq/ml in 3D mode. The noise equivalent count rate with 3D mode is equivalent to that with 2D mode at five times higher radioactivity level. The applicable imaging capabilities of the scanner was demonstrated by animal studies with a monkey.

  19. Design and performance evaluation of a high resolution IRI-microPET preclinical scanner

    NASA Astrophysics Data System (ADS)

    Islami rad, S. Z.; Peyvandi, R. Gholipour; lehdarboni, M. Askari; Ghafari, A. A.

    2015-05-01

    PET for small animal, IRI-microPET, was designed and built at the NSTRI. The scanner is made of four detectors positioned on a rotating gantry at a distance 50 mm from the center. Each detector consists of a 10×10 crystal matrix of 2×2×10 mm3 directly coupled to a PS-PMT. A position encoding circuit for specific PS-PMT has been designed, built and tested with a PD-MFS-2MS/s-8/14 data acquisition board. After implementing reconstruction algorithms (FBP, MLEM and SART) on sinograms, images quality and system performance were evaluated by energy resolution, timing resolution, spatial resolution, scatter fraction, sensitivity, RMS contrast and SNR parameters. The energy spectra were obtained for the crystals with an energy window of 300-700 keV. The energy resolution in 511 keV averaged over all modules, detectors, and crystals, was 23.5%. A timing resolution of 2.4 ns FWHM obtained by coincidence timing spectrum was measured with crystal LYSO. The radial and tangential resolutions for 18F (1.15-mm inner diameter) at the center of the field of view were 1.81 mm and 1.90 mm, respectively. At a radial offset of 5 mm, the FWHM values were 1.96 and 2.06 mm. The system scatter fraction was 7.1% for the mouse phantom. The sensitivity was measured for different energy windows, leading to a sensitivity of 1.74% at the center of FOV. Also, images quality was evaluated by RMS contrast and SNR factors, and the results show that the reconstructed images by MLEM algorithm have the best RMS contrast, and SNR. The IRI-microPET presents high image resolution, low scatter fraction values and improved SNR for animal studies.

  20. A count-rate model for PET scanners using pixelated Anger-logic detectors with different scintillators

    NASA Astrophysics Data System (ADS)

    Surti, S.; Karp, J. S.

    2005-12-01

    A high count-rate simulation (HCRSim) model has been developed so that all results are derived from fundamental physics principles. Originally developed to study the behaviour of continuous sodium iodide (NaI(Tl)) detectors, this model is now applied to PET scanners based on pixelated Anger-logic detectors using lanthanum bromide (LaBr3), gadolinium orthosilicate (GSO) and lutetium orthosilicate (LSO) scintillators. This simulation has been used to study the effect on scanner deadtime and pulse pileup at high activity levels due to the scintillator stopping power (μ), decay time (τ) and energy resolution. Simulations were performed for a uniform 20 cm diameter × 70 cm long cylinder (NEMA NU2-2001 standard) in a whole-body scanner with an 85 cm ring diameter and a 25 cm axial field-of-view. Our results for these whole-body scanners demonstrate the potential of a pixelated Anger-logic detector and the relationship of its performance with the scanner NEC rate. Faster signal decay and short coincidence timing window lead to a reduction in deadtime and randoms fraction in the LaBr3 and LSO scanners compared to GSO. The excellent energy resolution of LaBr3 leads to the lowest scatter fraction for all scanners and helps compensate for reduced sensitivity compared to the GSO and LSO scanners, leading to the highest NEC values at high activity concentrations. The LSO scanner has the highest sensitivity of all the scanner designs investigated here, therefore leading to the highest peak NEC value but at a lower activity concentration than that of LaBr3.

  1. A count-rate model for PET scanners using pixelated Anger-logic detectors with different scintillators.

    PubMed

    Surti, S; Karp, J S

    2005-12-01

    A high count-rate simulation (HCRSim) model has been developed so that all results are derived from fundamental physics principles. Originally developed to study the behaviour of continuous sodium iodide (NaI(Tl)) detectors, this model is now applied to PET scanners based on pixelated Anger-logic detectors using lanthanum bromide (LaBr(3)), gadolinium orthosilicate (GSO) and lutetium orthosilicate (LSO) scintillators. This simulation has been used to study the effect on scanner deadtime and pulse pileup at high activity levels due to the scintillator stopping power (mu), decay time (tau) and energy resolution. Simulations were performed for a uniform 20 cm diameter x 70 cm long cylinder (NEMA NU2-2001 standard) in a whole-body scanner with an 85 cm ring diameter and a 25 cm axial field-of-view. Our results for these whole-body scanners demonstrate the potential of a pixelated Anger-logic detector and the relationship of its performance with the scanner NEC rate. Faster signal decay and short coincidence timing window lead to a reduction in deadtime and randoms fraction in the LaBr(3) and LSO scanners compared to GSO. The excellent energy resolution of LaBr(3) leads to the lowest scatter fraction for all scanners and helps compensate for reduced sensitivity compared to the GSO and LSO scanners, leading to the highest NEC values at high activity concentrations. The LSO scanner has the highest sensitivity of all the scanner designs investigated here, therefore leading to the highest peak NEC value but at a lower activity concentration than that of LaBr(3). PMID:16306662

  2. Comparison of reconstruction methods and quantitative accuracy in Siemens Inveon PET scanner

    NASA Astrophysics Data System (ADS)

    Ram Yu, A.; Kim, Jin Su; Kang, Joo Hyun; Moo Lim, Sang

    2015-04-01

    PET reconstruction is key to the quantification of PET data. To our knowledge, no comparative study of reconstruction methods has been performed to date. In this study, we compared reconstruction methods with various filters in terms of their spatial resolution, non-uniformities (NU), recovery coefficients (RCs), and spillover ratios (SORs). In addition, the linearity of reconstructed radioactivity between linearity of measured and true concentrations were also assessed. A Siemens Inveon PET scanner was used in this study. Spatial resolution was measured with NEMA standard by using a 1 mm3 sized 18F point source. Image quality was assessed in terms of NU, RC and SOR. To measure the effect of reconstruction algorithms and filters, data was reconstructed using FBP, 3D reprojection algorithm (3DRP), ordered subset expectation maximization 2D (OSEM 2D), and maximum a posteriori (MAP) with various filters or smoothing factors (β). To assess the linearity of reconstructed radioactivity, image quality phantom filled with 18F was used using FBP, OSEM and MAP (β =1.5 & 5 × 10-5). The highest achievable volumetric resolution was 2.31 mm3 and the highest RCs were obtained when OSEM 2D was used. SOR was 4.87% for air and 3.97% for water, obtained OSEM 2D reconstruction was used. The measured radioactivity of reconstruction image was proportional to the injected one for radioactivity below 16 MBq/ml when FBP or OSEM 2D reconstruction methods were used. By contrast, when the MAP reconstruction method was used, activity of reconstruction image increased proportionally, regardless of the amount of injected radioactivity. When OSEM 2D or FBP were used, the measured radioactivity concentration was reduced by 53% compared with true injected radioactivity for radioactivity <16 MBq/ml. The OSEM 2D reconstruction method provides the highest achievable volumetric resolution and highest RC among all the tested methods and yields a linear relation between the measured and true

  3. Preliminary evaluation of a monolithic detector module for integrated PET/MRI scanner with high spatial resolution

    NASA Astrophysics Data System (ADS)

    Pani, R.; Gonzalez, A. J.; Bettiol, M.; Fabbri, A.; Cinti, M. N.; Preziosi, E.; Borrazzo, C.; Conde, P.; Pellegrini, R.; Di Castro, E.; Majewski, S.

    2015-06-01

    The proposal of Mindview European Project concerns with the development of a very high resolution and high efficiency brain dedicated PET scanner simultaneously working with a Magnetic Resonance scanner, that expects to visualize neurotransmitter pathways and their disruptions in the quest to better diagnose schizophrenia. On behalf of this project, we propose a low cost PET module for the first prototype, based on monolithic crystals, suitable to be integrated with a head Radio Frequency (RF) coil. The aim of the suggested module is to achieve high performances in terms of efficiency, planar spatial resolution (expected about 1 mm) and discrimination of gamma Depth Of Interaction (DOI) in order to reduce the parallax error. Our preliminary results are very promising: a DOI resolution of about 3 mm, a spatial resolution ranging from about 1 to 1.5 mm and a good position linearity.

  4. Progress in the development of a PET scanner based on BaF2 scintillator and photosensitive wire chambers.

    PubMed

    Tavernier, S; Bruyndonckx, P; Guerard, B; Zhang, S

    1991-01-01

    The work presented is part of a design study for a Positron Emission Tomograph (PET) scanner based on the use of BaF2 scintillator and photosensitive wire chambers. The detection efficiency for gamma radiation of 511 keV is found close to 100% for a sufficiently large crystal. For a matrix of small and elongated crystals as one would use in a PET scanner (5 x 5 x 50 mm3) we obtained 6 photoelectrons per 511 keV deposited. The following variants and alternatives were also studied: operation of the wire chamber at atmospheric pressure; double readout where the crystals are read on one side with a photomultiplier to give time and energy resolution, and on the other side with a wire chamber to localise the event; and Csl photocathodes. Encouraging results have been obtained for each of these, but particularly the Csl photocathodes look very promising. PMID:1839856

  5. Performance comparison of two commercial BGO-based PET/CT scanners using NEMA NU 2-2001

    SciTech Connect

    Bolard, Gregory; Prior, John O.; Modolo, Luca; Bischof Delaloye, Angelika; Kosinski, Marek; Wastiel, Claude; Malterre, Jerome; Bulling, Shelley; Bochud, Francois; Verdun, Francis R.

    2007-07-15

    Combined positron emission tomography and computed tomography (PET/CT) scanners play a major role in medicine for in vivo imaging in an increasing number of diseases in oncology, cardiology, neurology, and psychiatry. With the advent of short-lived radioisotopes other than {sup 18}F and newer scanners, there is a need to optimize radioisotope activity and acquisition protocols, as well as to compare scanner performances on an objective basis. The Discovery-LS (D-LS) was among the first clinical PET/CT scanners to be developed and has been extensively characterized with older National Electrical Manufacturer Association (NEMA) NU 2-1994 standards. At the time of publication of the latest version of the standards (NU 2-2001) that have been adapted for whole-body imaging under clinical conditions, more recent models from the same manufacturer, i.e., Discovery-ST (D-ST) and Discovery-STE (D-STE), were commercially available. We report on the full characterization both in the two- and three-dimensional acquisition mode of the D-LS according to latest NEMA NU 2-2001 standards (spatial resolution, sensitivity, count rate performance, accuracy of count losses, and random coincidence correction and image quality), as well as a detailed comparison with the newer D-ST widely used and whose characteristics are already published.

  6. Test of a single module of the J-PET scanner based on plastic scintillators

    NASA Astrophysics Data System (ADS)

    Moskal, P.; Niedźwiecki, Sz.; Bednarski, T.; Czerwiński, E.; Kapłon, Ł.; Kubicz, E.; Moskal, I.; Pawlik-Niedźwiecka, M.; Sharma, N. G.; Silarski, M.; Zieliński, M.; Zoń, N.; Białas, P.; Gajos, A.; Kochanowski, A.; Korcyl, G.; Kowal, J.; Kowalski, P.; Kozik, T.; Krzemień, W.; Molenda, M.; Pałka, M.; Raczyński, L.; Rudy, Z.; Salabura, P.; Słomski, A.; Smyrski, J.; Strzelecki, A.; Wieczorek, A.; Wiślicki, W.

    2014-11-01

    A Time of Flight Positron Emission Tomography scanner based on plastic scintillators is being developed at the Jagiellonian University by the J-PET collaboration. The main challenge of the conducted research lies in the elaboration of a method allowing application of plastic scintillators for the detection of low energy gamma quanta. In this paper we report on tests of a single detection module built out from the BC-420 plastic scintillator strip (with dimensions of 5×19×300 mm3) read out at two ends by Hamamatsu R5320 photomultipliers. The measurements were performed using collimated beam of annihilation quanta from the 68Ge isotope and applying the Serial Data Analyzer (Lecroy SDA6000A) which enabled sampling of signals with 50 ps intervals. The time resolution of the prototype module was established to be better than 80 ps (σ) for a single level discrimination. The spatial resolution of the determination of the hit position along the strip was determined to be about 0.93 cm (σ) for the annihilation quanta. The fractional energy resolution for the energy E deposited by the annihilation quanta via the Compton scattering amounts to σ(E) / E ≈ 0.044 /√{ E(MeV) } and corresponds to the σ(E) / E of 7.5% at the Compton edge.

  7. Monte Carlo simulation and scatter correction of the GE Advance PET scanner with SimSET and Geant4

    NASA Astrophysics Data System (ADS)

    Barret, Olivier; Carpenter, T. Adrian; Clark, John C.; Ansorge, Richard E.; Fryer, Tim D.

    2005-10-01

    For Monte Carlo simulations to be used as an alternative solution to perform scatter correction, accurate modelling of the scanner as well as speed is paramount. General-purpose Monte Carlo packages (Geant4, EGS, MCNP) allow a detailed description of the scanner but are not efficient at simulating voxel-based geometries (patient images). On the other hand, dedicated codes (SimSET, PETSIM) will perform well for voxel-based objects but will be poor in their capacity of simulating complex geometries such as a PET scanner. The approach adopted in this work was to couple a dedicated code (SimSET) with a general-purpose package (Geant4) to have the efficiency of the former and the capabilities of the latter. The combined SimSET+Geant4 code (SimG4) was assessed on the GE Advance PET scanner and compared to the use of SimSET only. A better description of the resolution and sensitivity of the scanner and of the scatter fraction was obtained with SimG4. The accuracy of scatter correction performed with SimG4 and SimSET was also assessed from data acquired with the 20 cm NEMA phantom. SimG4 was found to outperform SimSET and to give slightly better results than the GE scatter correction methods installed on the Advance scanner (curve fitting and scatter modelling for the 300-650 keV and 375-650 keV energy windows, respectively). In the presence of a hot source close to the edge of the field of view (as found in oxygen scans), the GE curve-fitting method was found to fail whereas SimG4 maintained its performance.

  8. Monte Carlo simulation and scatter correction of the GE advance PET scanner with SimSET and Geant4.

    PubMed

    Barret, Olivier; Carpenter, T Adrian; Clark, John C; Ansorge, Richard E; Fryer, Tim D

    2005-10-21

    For Monte Carlo simulations to be used as an alternative solution to perform scatter correction, accurate modelling of the scanner as well as speed is paramount. General-purpose Monte Carlo packages (Geant4, EGS, MCNP) allow a detailed description of the scanner but are not efficient at simulating voxel-based geometries (patient images). On the other hand, dedicated codes (SimSET, PETSIM) will perform well for voxel-based objects but will be poor in their capacity of simulating complex geometries such as a PET scanner. The approach adopted in this work was to couple a dedicated code (SimSET) with a general-purpose package (Geant4) to have the efficiency of the former and the capabilities of the latter. The combined SimSET+Geant4 code (SimG4) was assessed on the GE Advance PET scanner and compared to the use of SimSET only. A better description of the resolution and sensitivity of the scanner and of the scatter fraction was obtained with SimG4. The accuracy of scatter correction performed with SimG4 and SimSET was also assessed from data acquired with the 20 cm NEMA phantom. SimG4 was found to outperform SimSET and to give slightly better results than the GE scatter correction methods installed on the Advance scanner (curve fitting and scatter modelling for the 300-650 keV and 375-650 keV energy windows, respectively). In the presence of a hot source close to the edge of the field of view (as found in oxygen scans), the GE curve-fitting method was found to fail whereas SimG4 maintained its performance. PMID:16204875

  9. Voxelwise lp-ntPET for detecting localized, transient dopamine release of unknown timing: Sensitivity Analysis and Application to Cigarette Smoking in the PET Scanner

    PubMed Central

    Kim, Su Jin; Sullivan, Jenna M; Wang, Shuo; Cosgrove, Kelly P; Morris, Evan D

    2014-01-01

    The “linear parametric neurotransmitter PET” (lp-ntPET) model estimates time variation in endogenous neurotransmitter levels from dynamic PET data. The pattern of dopamine (DA) change over time may be an important element of the brain's response to addictive substances such as cigarettes or alcohol. We have extended the lp-ntPET model from the original region of interest (ROI) - based implementation to be able to apply the model at the voxel level. The resulting endpoint is a dynamic image, or movie, of transient neurotransmitter changes. Simulations were performed to select threshold values to reduce the false positive rate when applied to real 11C-raclopride PET data. We tested the new voxelwise method on simulated data, and finally, we applied it to 11C-raclopride PET data of subjects smoking cigarettes in the PET scanner. In simulation, the temporal precision of neurotransmitter response was shown to be similar to that of ROI-based lp-ntPET (standard deviation ∼ 3 min). False positive rates for the voxelwise method were well controlled by combining a statistical threshold (the F-test) with a new spatial (cluster-size) thresholding operation. Sensitivity of detection for the new algorithm was greater than 80% for the case of short-lived DA changes that occur in subregions of the striatum as might be the case with cigarette smoking. Finally, in 11C-raclopride PET data, DA movies reveal for the first time that different temporal patterns of the DA response to smoking may exist in different subregions of the striatum. These spatiotemporal patterns of neurotransmitter change created by voxelwise lp-ntPET may serve as novel biomarkers for addiction and/or treatment efficacy. PMID:24700424

  10. Performance evaluation of the new whole-body PET/CT scanner: Discovery ST.

    PubMed

    Bettinardi, Valentino; Danna, Massimo; Savi, Annarita; Lecchi, Michela; Castiglioni, Isabella; Gilardi, Maria Carla; Bammer, Helmut; Lucignani, Giovanni; Fazio, Ferruccio

    2004-06-01

    Characterisation of the physical performance of the new integrated PET/CT system Discovery ST (GE Medical Systems) has been performed following the NEMA NU 2-1994 (N-94) and the NEMA NU 2-2001 (N-01) standards in both 2D and 3D acquisition configuration. The Discovery ST combines a four or eight multi-slice helical CT scanner with a PET tomograph which consists of 10,080 BGO crystals arranged in 24 rings. The crystal dimensions are 6.3 x 6.3 x 30 mm(3) and they are organised in blocks of 6 x 6 crystals, coupled to a single photomultiplier tube with four anodes. The 24 rings of the PET system allow 47 images to be obtained, spaced by 3.27 mm, and covering an axial field of view of 157 mm. The low- and high-energy thresholds are set to 375 and 650 keV, respectively. The coincidence time window is set to 11.7 ns. Using the NEMA N-94 standard, the main results were: (1) the average (radial and tangential) transverse spatial resolution (FWHM) at 1, 10 and 20 cm off axis was 6.28 mm, 7.09 mm and 7.45 mm in 2D, and 6.68 mm, 7.72 mm and 8.13 mm in 3D; (2) the sensitivity for true events was 8,567 cps/kBq/cc in 2D and 36,649 cps/kBq/cc in 3D; (3) the scatter fraction was 15% in 2D and 30% in 3D; (4) the peak true events rate, the true events rate at 50% of the system dead-time and the true events rate when equal to the random events rate were 750 kcps at 189.81 kBq/cc, 744 kcps at 186.48 kBq/cc and 686 kcps at 150.59 kBq/cc, respectively, in 2D, and 922 kcps at 44.03 kBq/cc, 834 kcps at 53.28 kBq/cc and 921 kcps at 44.03 kBq/cc in 3D; (5) the noise equivalent count (NEC) peak rate was 270 kcps at 34.38 kBq/cc in 3D, with random coincidences estimated by delayed events. Using the NEMA N-01 standards the main results were: (1) the average transverse and axial spatial resolution (FWHM) at 1 cm and 10 cm off axis was 6.28 (4.56) mm and 6.88 (6.11) mm in 2D, and 6.29 (5.68) mm and 6.82 (6.05) mm in 3D; (2) the average sensitivity for the two radial positions (r=0 cm and r=10 cm

  11. Ga-68-DOTATOC: Feasibility of high throughput screening by small animal PET using a clinical high-resolution PET/CT scanner

    NASA Astrophysics Data System (ADS)

    Hofmann, Michael; Weitzel, Thilo; Krause, Thomas

    2006-12-01

    As radio peptide tracers have been developed in recent years for the high sensitive detection of neuroendocrine tumors, still the broad application of other peptides to breast and prostate cancer is missing. A rapid screening of new peptides can, in theory, be based on in vivo screening in animals by PET/CT. To test this hypothesis and to asses the minimum screening time needed per animal, we used the application of Ga-68-DOTATOC PET/CT in rats as test system. The Ga-68-DOTATOC yields in a hot spot imaging with minimal background. To delineate liver and spleen, we performed PET/CT of 10 animals on a SIEMENS Biograph 16 LSO HIGHREZ after intravenous injection of 1.5 MBq Ga-68-DOTATOC per animal. Animals were mounted in an '18 slot' holding device and scanned for a single-bed position. The emission times for the PET scan was varied from 1 to 20 min. The images were assessed first for "PET only" and afterwards in PET/CT fusion mode. The detection of the two organs was good at emission times down to 1 min in PET/CT fusion mode. In the "PET only" scans, the liver was clearly to be identified down to 1 min emission in all animals. But the spleen could only be delineated only by 1 min of emission in the PET/CT-fusion mode. In conclusion the screening of "hot spot" enriching peptides is feasible. "PET only" is in terms of delineation of small organs by far inferior to PET/CT fusion. If animal tumors are above a diameter of 10 mm small, animal PET/CT using clinical high resolution scanners will enable rapid screening. Even the determination of bio-distributions becomes feasible by using list mode tools. The time for the whole survey of 18 animals including anesthesia, preparation and mounting was approximately 20 min. By use of several holding devices mounted simultaneously, a survey time of less than 1 h for 180 animals can be expected.

  12. SiliPET: Design of an ultra-high resolution small animal PET scanner based on stacks of semi-conductor detectors

    NASA Astrophysics Data System (ADS)

    Cesca, N.; Auricchio, N.; Di Domenico, G.; Zavattini, G.; Malaguti, R.; Andritschke, R.; Kanbach, G.; Schopper, F.

    2007-03-01

    We studied with Monte Carlo simulations, using the EGSnrc code, a new scanner for small animal positron emission tomography (PET), based on stacks of double-sided semiconductor detectors. Each stack is composed of planar detectors with dimension 70×60×1 mm 3 and orthogonal strips on both sides with 500 μm pitch to read the two interaction coordinates, the third being the detector number in the stack. Multiple interactions in a stack are discarded. In this way, we achieve a precise determination of the first interaction point of the two 511 keV photons. The reduced dimensions of the scanner also improve the solid angle coverage resulting in a high sensitivity. Preliminary results of scanners based on Si planar detectors are presented and the initial tomographic reconstructions demonstrate very good spatial resolution limited only by the positron range. This suggests that, this is a promising new approach for small animal PET imaging. We are testing some double-sided silicon detectors, equipped with 128 orthogonal p and n strips on opposite sides using VATAGP3 ASIC by IDEAS.

  13. Simulation of the Expected Performance of a Seamless Scanner for Brain PET Based on Highly Pixelated CdTe Detectors

    PubMed Central

    Mikhaylova, Ekaterina; De Lorenzo, Gianluca; Chmeissani, Mokhtar; Kolstein, Machiel; Cañadas, Mario; Arce, Pedro; Calderón, Yonatan; Uzun, Dilber; Ariño, Gerard; Macias-Montero, José Gabriel; Martinez, Ricardo; Puigdengoles, Carles; Cabruja, Enric

    2014-01-01

    The aim of this work is the evaluation of the design for a nonconventional PET scanner, the voxel imaging PET (VIP), based on pixelated room-temperature CdTe detectors yielding a true 3-D impact point with a density of 450 channels cm3, for a total 6 336 000 channels in a seamless ring shaped volume. The system is simulated and evaluated following the prescriptions of the NEMA NU 2-2001 and the NEMA NU 4-2008 standards. Results show that the excellent energy resolution of the CdTe detectors (1.6% for 511 keV photons), together with the small voxel pitch (1×1×2 mm3), and the crack-free ring geometry, give the design the potential to overcome the current limitations of PET scanners and to approach the intrinsic image resolution limits set by physics. The VIP is expected to reach a competitive sensitivity and a superior signal purity with respect to values commonly quoted for state-of-the-art scintillating crystal PETs. The system can provide 14 cps/kBq with a scatter fraction of 3.95% and 21 cps/kBq with a scatter fraction of 0.73% according to NEMA NU 2-2001 and NEMA NU 4-2008, respectively. The calculated NEC curve has a peak value of 122 kcps at 5.3 kBq/mL for NEMA NU 2-2001 and 908 kcps at 1.6 MBq/mL for NEMA NU 4-2008. The proposed scanner can achieve an image resolution of ~ 1 mm full-width at half-maximum in all directions. The virtually noise-free data sample leads to direct positive impact on the quality of the reconstructed images. As a consequence, high-quality high-resolution images can be obtained with significantly lower number of events compared to conventional scanners. Overall, simulation results suggest the VIP scanner can be operated either at normal dose for fast scanning and high patient throughput, or at low dose to decrease the patient radioactivity exposure. The design evaluation presented in this work is driving the development and the optimization of a fully operative prototype to prove the feasibility of the VIP concept. PMID:24108750

  14. Synthesis and quality control of fluorodeoxyglucose and performance assessment of Siemens MicroFocus 220 small animal PET scanner

    NASA Astrophysics Data System (ADS)

    Phaterpekar, Siddhesh Nitin

    The scope of this article is to cover the synthesis and quality control procedures involved in production of Fludeoxyglucose (18F--FDG). The article also describes the cyclotron production of 18F radioisotope and gives a brief overview on operations and working of a fixed energy medical cyclotron. The quality control procedures for FDG involve radiochemical and radionuclidic purity tests, pH tests, chemical purity tests, sterility tests, endotoxin tests. Each of these procedures were carried out for multiple batches of FDG with a passing rate of 95% among 20 batches. The article also covers the quality assurance steps for the Siemens MicroPET Focus 220 Scanner using a Jaszczak phantom. We have carried out spatial resolution tests on the scanner, with an average transaxial resolution of 1.775mm with 2-3mm offset. Tests involved detector efficiency, blank scan sinograms and transmission sinograms. A series of radioactivity distribution tests are also carried out on a uniform phantom, denoting the variations in radioactivity and uniformity by using cylindrical ROIs in the transverse region of the final image. The purpose of these quality control tests is to make sure the manufactured FDG is biocompatible with the human body. Quality assurance tests are carried on PET scanners for efficient performance, and to make sure the quality of images acquired is according to the radioactivity distribution in the subject of interest.

  15. Physical and clinical performance of the mCT time-of-flight PET/CT scanner

    NASA Astrophysics Data System (ADS)

    Jakoby, B. W.; Bercier, Y.; Conti, M.; Casey, M. E.; Bendriem, B.; Townsend, D. W.

    2011-04-01

    Time-of-flight (TOF) measurement capability promises to improve PET image quality. We characterized the physical and clinical PET performance of the first Biograph mCT TOF PET/CT scanner (Siemens Medical Solutions USA, Inc.) in comparison with its predecessor, the Biograph TruePoint TrueV. In particular, we defined the improvements with TOF. The physical performance was evaluated according to the National Electrical Manufacturers Association (NEMA) NU 2-2007 standard with additional measurements to specifically address the TOF capability. Patient data were analyzed to obtain the clinical performance of the scanner. As expected for the same size crystal detectors, a similar spatial resolution was measured on the mCT as on the TruePoint TrueV. The mCT demonstrated modestly higher sensitivity (increase by 19.7 ± 2.8%) and peak noise equivalent count rate (NECR) (increase by 15.5 ± 5.7%) with similar scatter fractions. The energy, time and spatial resolutions for a varying single count rate of up to 55 Mcps resulted in 11.5 ± 0.2% (FWHM), 527.5 ± 4.9 ps (FWHM) and 4.1 ± 0.0 mm (FWHM), respectively. With the addition of TOF, the mCT also produced substantially higher image contrast recovery and signal-to-noise ratios in a clinically-relevant phantom geometry. The benefits of TOF were clearly demonstrated in representative patient images.

  16. Performance evaluation of the Ingenuity TF PET/CT scanner with a focus on high count-rate conditions

    PubMed Central

    Kolthammer, Jeffrey A; Su, Kuan-Hao; Grover, Anu; Narayanan, Manoj; Jordan, David W; Muzic, Raymond F

    2015-01-01

    This study evaluated the positron emission tomography (PET) imaging performance of the Ingenuity TF 128 PET/computed tomography (CT) scanner which has a PET component that was designed to support a wider radioactivity range than is possible with those of Gemini TF PET/CT and Ingenuity TF PET/MR. Spatial resolution, sensitivity, count rate characteristics and image quality were evaluated according to the NEMA NU 2–2007 standard and ACR phantom accreditation procedures; these were supplemented by additional measurements intended to characterize the system under conditions that would be encountered during quantitative cardiac imaging with 82Rb. Image quality was evaluated using a hot spheres phantom, and various contrast recovery and noise measurements were made from replicated images. Timing and energy resolution, dead time, and the linearity of the image activity concentration, were all measured over a wide range of count rates. Spatial resolution (4.8– 5.1 mm FWHM), sensitivity (7.3 cps kBq−1), peak noise-equivalent count rate (124 kcps), and peak trues rate (365 kcps)were similar to those of the Gemini TF PET/CT. Contrast recovery was higher with a 2 mm, body-detail reconstruction than with a 4 mm, body reconstruction, although the precision was reduced. The noise equivalent count rate peak was broad (within 10% of peak from 241–609 MBq). The activity measured in phantom images was within 10% of the true activity for count rates up to those observed in 82Rb cardiac PET studies. PMID:24955921

  17. An OpenPET scanner with bridged detectors to compensate for incomplete data.

    PubMed

    Tashima, Hideaki; Yamaya, Taiga; Kinahan, Paul E

    2014-10-21

    We are developing an open-type PET 'OpenPET' geometry. One possible geometry is a dual-ring OpenPET, which consists of two detector rings separated by a gap for entrance of a radiotherapy beam or for inserting other modalities. In our previous simulations and experiments the OpenPET imaging geometry was shown to be feasible by applying iterative reconstruction methods. However, the gap violates Orlov's completeness condition for accurate tomographic reconstruction. In this study, we propose a solution for the incompleteness problem by adding bridge detectors to fill in parts of the gaps of the OpenPET geometry; we call this bridged OpenPET. Although this geometry was considered previously, its analytical property was not discussed. Therefore, we applied the direct Fourier method as an analytical reconstruction method to the bridged OpenPET, dual-ring OpenPET and conventional cylindrical PET for comparison. Numerical simulations showed that the additional bridge detectors compensate for the incompleteness of the OpenPET by covering one direction perpendicular to the transaxial slices of the imaging subjects. PMID:25255296

  18. An OpenPET scanner with bridged detectors to compensate for incomplete data

    NASA Astrophysics Data System (ADS)

    Tashima, Hideaki; Yamaya, Taiga; Kinahan, Paul E.

    2014-10-01

    We are developing an open-type PET ‘OpenPET’ geometry. One possible geometry is a dual-ring OpenPET, which consists of two detector rings separated by a gap for entrance of a radiotherapy beam or for inserting other modalities. In our previous simulations and experiments the OpenPET imaging geometry was shown to be feasible by applying iterative reconstruction methods. However, the gap violates Orlov’s completeness condition for accurate tomographic reconstruction. In this study, we propose a solution for the incompleteness problem by adding bridge detectors to fill in parts of the gaps of the OpenPET geometry; we call this bridged OpenPET. Although this geometry was considered previously, its analytical property was not discussed. Therefore, we applied the direct Fourier method as an analytical reconstruction method to the bridged OpenPET, dual-ring OpenPET and conventional cylindrical PET for comparison. Numerical simulations showed that the additional bridge detectors compensate for the incompleteness of the OpenPET by covering one direction perpendicular to the transaxial slices of the imaging subjects.

  19. MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging

    NASA Astrophysics Data System (ADS)

    Tai, Yuan-Chuan; Chatziioannou, Arion F.; Yang, Yongfeng; Silverman, Robert W.; Meadors, Ken; Siegel, Stefan; Newport, Danny F.; Stickel, Jennifer R.; Cherry, Simon R.

    2003-06-01

    MicroPET II is a second-generation animal PET scanner designed for high-resolution imaging of small laboratory rodents. The system consists of 90 scintillation detector modules arranged in three contiguous axial rings with a ring diameter of 16.0 cm and an axial length of 4.9 cm. Each detector module consists of a 14 × 14 array of lutetium oxyorthosilicate (LSO) crystals coupled to a multi-channel photomultiplier tube (MC-PMT) through a coherent optical fibre bundle. Each LSO crystal element measures 0.975 mm × 0.975 mm in cross section by 12.5 mm in length. A barium sulphate reflector material was used between LSO elements leading to a detector pitch of 1.15 mm in both axial and transverse directions. Fused optical fibre bundles were made from 90 µm diameter glass fibres with a numerical aperture of 0.56. Interstitial extramural absorber was added between the fibres to reduce optical cross talk. A charge-division readout circuit was implemented on printed circuit boards to decode the 196 crystals in each array from the outputs of the 64 anode signals of the MC-PMT. Electronics from Concorde Microsystems Inc. (Knoxville, TN) were used for signal amplification, digitization, event qualification, coincidence processing and data capture. Coincidence data were passed to a host PC that recorded events in list mode. Following acquisition, data were sorted into sinograms and reconstructed using Fourier rebinning and filtered backprojection algorithms. Basic evaluation of the system has been completed. The absolute sensitivity of the microPET II scanner was 2.26% at the centre of the field of view (CFOV) for an energy window of 250-750 keV and a timing window of 10 ns. The intrinsic spatial resolution of the detectors in the system averaged 1.21 mm full width at half maximum (FWHM) when measured with a 22Na point source 0.5 mm in diameter. Reconstructed image resolution ranged from 0.83 mm FWHM at the CFOV to 1.47 mm FWHM in the radial direction, 1.17 mm FWHM in the

  20. MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging.

    PubMed

    Tai, Yuan-Chuan; Chatziioannou, Arion F; Yang, Yongfeng; Silverman, Robert W; Meadors, Ken; Siegel, Stefan; Newport, Danny F; Stickel, Jennifer R; Cherry, Simon R

    2003-06-01

    MicroPET II is a second-generation animal PET scanner designed for high-resolution imaging of small laboratory rodents. The system consists of 90 scintillation detector modules arranged in three contiguous axial rings with a ring diameter of 16.0 cm and an axial length of 4.9 cm. Each detector module consists of a 14 x 14 array of lutetium oxyorthosilicate (LSO) crystals coupled to a multi-channel photomultiplier tube (MC-PMT) through a coherent optical fibre bundle. Each LSO crystal element measures 0.975 mm x 0.975 mm in cross section by 12.5 mm in length. A barium sulphate reflector material was used between LSO elements leading to a detector pitch of 1.15 mm in both axial and transverse directions. Fused optical fibre bundles were made from 90 microm diameter glass fibres with a numerical aperture of 0.56. Interstitial extramural absorber was added between the fibres to reduce optical cross talk. A charge-division readout circuit was implemented on printed circuit boards to decode the 196 crystals in each array from the outputs of the 64 anode signals of the MC-PMT. Electronics from Concorde Microsystems Inc. (Knoxville, TN) were used for signal amplification, digitization, event qualification, coincidence processing and data capture. Coincidence data were passed to a host PC that recorded events in list mode. Following acquisition, data were sorted into sinograms and reconstructed using Fourier rebinning and filtered hackprojection algorithms. Basic evaluation of the system has been completed. The absolute sensitivity of the microPET II scanner was 2.26% at the centre of the field of view (CFOV) for an energy window of 250-750 keV and a timing window of 10 ns. The intrinsic spatial resolution of the detectors in the system averaged 1.21 mm full width at half maximum (FWHM) when measured with a 22Na point source 0.5 mm in diameter. Reconstructed image resolution ranged from 0.83 mm FWHM at the CFOV to 1.47 mm FWHM in the radial direction, 1.17 mm FWHM in the

  1. Development and evaluation of an ultra-fast ASIC for future PET scanners using TOF-capable MPPC array detectors

    NASA Astrophysics Data System (ADS)

    Ambe, T.; Ikeda, H.; Kataoka, J.; Matsuda, H.; Kato, T.

    2015-01-01

    We developed a front-end ASIC for future PET scanners with Time-Of-Flight (TOF) capability to be coupled with 4×4 Multi-Pixel Photon Counter (MPPC) arrays. The ASIC is designed based on the open-IP project proposed by JAXA and realized in TSMC 0.35 μm CMOS technology. The circuit comprises 16-channel, low impedance current conveyors for effectively acquiring fast MPPC signals. For precise measurement of the coincidence timing of 511-keV gamma rays, the leading-edge method was used to discriminate the signals. We first tested the time response of the ASIC by illuminating each channel of a MPPC array device 3×3 mm2 in size with a Pico-second Light Pulsar with a light emission peak of 655 nm and pulse duration of 54 ps (FWHM). We obtained 105 ps (FWHM) on average for each channel in time jitter measurements. Moreover, we compensated for the time lag of each channel with inner delay circuits and succeeded in suppressing about a 700-ps lag to only 15 ps. This paper reports TOF measurements using back-to-back 511-keV signals, and suggests that the ASIC can be a promising device for future TOF-PET scanners based on the MPPC array.

  2. New shielding configurations for a simultaneous PET/MRI scanner at 7T.

    PubMed

    Peng, Bo J; Wu, Yibao; Cherry, Simon R; Walton, Jeffrey H

    2014-02-01

    Understanding sources of electromagnetic interference are important in designing any electronic system. This is especially true when combining positron emission tomography (PET) and magnetic resonance imaging (MRI) in a multimodality system as coupling between the subsystems can degrade the performance of either modality. For this reason, eliminating radio frequency (RF) interference and gradient-induced eddy currents have been major challenges in building simultaneous hybrid PET/MRI systems. MRI requires negligible RF interference at the Larmor resonance frequency, while RF interference at almost any frequency may corrupt PET data. Moreover, any scheme that minimizes these interactions would, ideally, not compromise the performance of either subsystem. This paper lays out a plan to resolve these problems. A carbon fiber composite material is found to be a good RF shield at the Larmor frequency (300MHz in this work) while introducing negligible gradient eddy currents. This carbon fiber composite also provides excellent structural support for the PET detector components. Low frequency electromagnetic radiation (81kHz here) from the switching power supplies of the gradient amplifiers was also found to interfere with the PET detector. Placing the PET detector module between two carbon fiber tubes and grounding the inner carbon fiber tube to the PET detector module ground reduced this interference. Further reductions were achieved by adding thin copper (Cu) foil on the outer carbon fiber case and electrically grounding the PET detector module so that all 3 components had a common ground, i.e. with the PET detector in an electrostatic cage. Finally, gradient switching typical in MRI sequences can result in count losses in the particular PET detector design studied. Moreover, the magnitude of this effect depends on the location of the detector within the magnet bore and which MRI gradient is being switched. These findings have a bearing on future designs of PET

  3. New shielding configurations for a simultaneous PET/MRI scanner at 7T

    NASA Astrophysics Data System (ADS)

    Peng, Bo J.; Wu, Yibao; Cherry, Simon R.; Walton, Jeffrey H.

    2014-02-01

    Understanding sources of electromagnetic interference are important in designing any electronic system. This is especially true when combining positron emission tomography (PET) and magnetic resonance imaging (MRI) in a multimodality system as coupling between the subsystems can degrade the performance of either modality. For this reason, eliminating radio frequency (RF) interference and gradient-induced eddy currents have been major challenges in building simultaneous hybrid PET/MRI systems. MRI requires negligible RF interference at the Larmor resonance frequency, while RF interference at almost any frequency may corrupt PET data. Moreover, any scheme that minimizes these interactions would, ideally, not compromise the performance of either subsystem. This paper lays out a plan to resolve these problems. A carbon fiber composite material is found to be a good RF shield at the Larmor frequency (300 MHz in this work) while introducing negligible gradient eddy currents. This carbon fiber composite also provides excellent structural support for the PET detector components. Low frequency electromagnetic radiation (81 kHz here) from the switching power supplies of the gradient amplifiers was also found to interfere with the PET detector. Placing the PET detector module between two carbon fiber tubes and grounding the inner carbon fiber tube to the PET detector module ground reduced this interference. Further reductions were achieved by adding thin copper (Cu) foil on the outer carbon fiber case and electrically grounding the PET detector module so that all 3 components had a common ground, i.e. with the PET detector in an electrostatic cage. Finally, gradient switching typical in MRI sequences can result in count losses in the particular PET detector design studied. Moreover, the magnitude of this effect depends on the location of the detector within the magnet bore and which MRI gradient is being switched. These findings have a bearing on future designs of PET

  4. Development of a MPPC-based prototype gantry for future MRI-PET scanners

    NASA Astrophysics Data System (ADS)

    Kurei, Y.; Kataoka, J.; Kato, T.; Fujita, T.; Ohshima, T.; Taya, T.; Yamamoto, S.

    2014-12-01

    We have developed a high spatial resolution, compact Positron Emission Tomography (PET) module designed for small animals and intended for use in magnetic resonance imaging (MRI) systems. This module consists of large-area, 4 × 4 ch MPPC arrays (S11830-3344MF; Hamamatsu Photonics K.K.) optically coupled with Ce-doped (Lu,Y)2(SiO4)O (Ce:LYSO) scintillators fabricated into 16 × 16 matrices of 0.5 × 0.5 mm2 pixels. We set the temperature sensor (LM73CIMK-0; National Semiconductor Corp.) at the rear of the MPPC acceptance surface, and apply optimum voltage to maintain the gain. The eight MPPC-based PET modules and coincidence circuits were assembled into a gantry arranged in a ring 90 mm in diameter to form the MPPC-based PET system. We have developed two types PET gantry: one made of non-magnetic metal and the other made of acrylonitrile butadiene styrene (ABS) resins. The PET gantry was positioned around the RF coil of the 4.7 T MRI system. We took an image of a point }22Na source under fast spin echo (FSE) and gradient echo (GE), in order to measure the interference between the MPPC-based PET and MRI. The spatial resolution of PET imaging in a transaxial plane of about 1 mm (FWHM) was achieved in all cases. Operating with PET made of ABS has no effect on MR images, while operating with PET made of non-magnetic metal has a significant detrimental effect on MR images. This paper describes our quantitative evaluations of PET images and MR images, and presents a more advanced version of the gantry for future MRI/DOI-PET systems.

  5. Evaluation of the ECAT EXACT HR{sup +} 3D PET scanner in {sup 15}O-water brain activation studies

    SciTech Connect

    Moreno-Cantu, J.J.; Thompson, C.J.; Zatorre, R.J.

    1996-12-31

    We evaluated the performance of the ECAT EXACT HR{sup +} 3D whole body PET scanner when employed to measure brain function using {sup 15}O-water-bolus activation protocols in single data acquisition sessions. Using vibrotactile and auditory stimuli as independent activation tasks, we studied the scanner`s performance under different imaging conditions in four healthy volunteers. Cerebral blood flow images were acquired from each volunteer using {sup 15}O-water-bolus injections of activity varying from 5 to 20mCi. Performance characteristics. The scanner`s dead time grew linearly with injected dose from 10% to 25%. Random events varied from 30% to 50% of the detected events. Scattered events were efficiently corrected at all doses. Noise-effective-count curves plateau at about 15mCi. One-session 12-injection bolus PET activation protocol. Using an acquisition protocol that accounts for the scanner`s performance and the practical aspects of imaging volunteers and patients in one session, we assessed the correlation between the statistical significance of activation foci and the dose per injection used The one-session protocol employs 12 bolus injections per subject. We present evidence suggesting that 15-20mCi is the optimal dose per injection to be used routinely in one-time scanning sessions.

  6. A novel adaptive discrete cosine transform-domain filter for gap-inpainting of high resolution PET scanners

    SciTech Connect

    Shih, Cheng-Ting; Lin, Hsin-Hon; Chuang, Keh-Shih; Wu, Jay; Chang, Shu-Jun

    2014-08-15

    Purpose: Several positron emission tomography (PET) scanners with special detector block arrangements have been developed in recent years to improve the resolution of PET images. However, the discontinuous detector blocks cause gaps in the sinogram. This study proposes an adaptive discrete cosine transform-based (aDCT) filter for gap-inpainting. Methods: The gap-corrupted sinogram was morphologically closed and subsequently converted to the DCT domain. A certain number of the largest coefficients in the DCT spectrum were identified to determine the low-frequency preservation region. The weighting factors for the remaining coefficients were determined by an exponential weighting function. The aDCT filter was constructed and applied to two digital phantoms and a simulated phantom introduced with various levels of noise. Results: For the Shepp-Logan head phantom, the aDCT filter filled the gaps effectively. For the Jaszczak phantom, no secondary artifacts were induced after aDCT filtering. The percent mean square error and mean structure similarity of the aDCT filter were superior to those of the DCT2 filter at all noise levels. For the simulated striatal dopamine innervation study, the aDCT filter recovered the shape of the striatum and restored the striatum to reference activity ratios to the ideal value. Conclusions: The proposed aDCT filter can recover the missing gap data in the sinogram and improve the image quality and quantitative accuracy of PET images.

  7. Personnel radiation dose considerations in the use of an integrated PET-CT scanner for radiotherapy treatment planning.

    PubMed

    Carson, K J; Young, V A L; Cosgrove, V P; Jarritt, P H; Hounsell, A R

    2009-11-01

    The acquisition of radiotherapy planning scans on positron emission tomography (PET)-CT scanners requires the involvement of radiotherapy radiographers. This study assessed the radiation dose received by these radiographers during this process. Radiotherapy planning (18)F-fluorodeoxyglucose ((18)F-FDG) PET-CT scans were acquired for 28 non-small cell lung cancer patients. In order to minimise the radiation dose received, a two-stage process was used in which the most time-consuming part of the set-up was performed before the patient received their (18)F-FDG injection. Throughout this process, the radiographers wore electronic personal dosemeters and recorded the doses received at different stages of the process. The mean total radiation dose received by a radiotherapy radiographer was 5.1+/-2.6 microSv per patient. The use of the two-stage process reduced the time spent in close proximity to the patient by approximately a factor of four. The two-stage process was effective in keeping radiation dose to a minimum. The use of a pre-injection set-up session reduces the radiation dose to the radiotherapy radiographers because of their involvement in PET-CT radiotherapy treatment planning scans by approximately a factor of three. PMID:19332513

  8. Impact of detector design on imaging performance of a long axial field-of-view, whole-body PET scanner

    PubMed Central

    Surti, S; Karp, J S

    2015-01-01

    Current generation of commercial time-of-flight (TOF) PET scanners utilize 20–25 mm thick LSO or LYSO crystals and have an axial FOV (AFOV) in the range of 16–22 mm. Longer AFOV scanners would provide increased intrinsic sensitivity and require fewer bed positions for whole-body imaging. Recent simulation work has investigated the sensitivity gains that can be achieved with these long AFOV scanners, and has motivated new areas of investigation such as imaging with very low dose of injected activity as well as providing whole-body dynamic imaging capability in one bed position. In this simulation work we model a 72 cm long scanner and prioritize the detector design choices in terms of timing resolution, crystal size (spatial resolution), crystal thickness (detector sensitivity), and depth-of-interaction (DOI) measurement capability. The generated list data are reconstructed with a list-mode OSEM algorithm using a Gaussian TOF kernel that depends on the timing resolution and blob basis functions for regularization. We use lesion phantoms and clinically relevant metrics for lesion detectability and contrast measurement. The scan time was fixed at 10 minutes for imaging a 100 cm long object assuming a 50% overlap between adjacent bed positions. Results show that a 72 cm long scanner can provide a factor of ten reduction in injected activity compared to an identical 18 cm long scanner to get equivalent lesion detectability. While improved timing resolution leads to further gains, using 3 mm (as opposed to 4 mm) wide crystals does not show any significant benefits for lesion detectability. A detector providing 2-level DOI information with equal crystal thickness also does not show significant gains. Finally, a 15 mm thick crystal leads to lower lesion detectability than a 20 mm thick crystal when keeping all other detector parameters (crystal width, timing resolution, and DOI capability) the same. However, improved timing performance with 15 mm thick crystals can

  9. Impact of detector design on imaging performance of a long axial field-of-view, whole-body PET scanner

    NASA Astrophysics Data System (ADS)

    Surti, S.; Karp, J. S.

    2015-07-01

    Current generation of commercial time-of-flight (TOF) PET scanners utilize 20-25 mm thick LSO or LYSO crystals and have an axial FOV (AFOV) in the range of 16-22 mm. Longer AFOV scanners would provide increased intrinsic sensitivity and require fewer bed positions for whole-body imaging. Recent simulation work has investigated the sensitivity gains that can be achieved with these long AFOV scanners, and has motivated new areas of investigation such as imaging with a very low dose of injected activity as well as providing whole-body dynamic imaging capability in one bed position. In this simulation work we model a 72 cm long scanner and prioritize the detector design choices in terms of timing resolution, crystal size (spatial resolution), crystal thickness (detector sensitivity), and depth-of-interaction (DOI) measurement capability. The generated list data are reconstructed with a list-mode OSEM algorithm using a Gaussian TOF kernel that depends on the timing resolution and blob basis functions for regularization. We use lesion phantoms and clinically relevant metrics for lesion detectability and contrast measurement. The scan time was fixed at 10 min for imaging a 100 cm long object assuming a 50% overlap between adjacent bed positions. Results show that a 72 cm long scanner can provide a factor of ten reduction in injected activity compared to an identical 18 cm long scanner to get equivalent lesion detectability. While improved timing resolution leads to further gains, using 3 mm (as opposed to 4 mm) wide crystals does not show any significant benefits for lesion detectability. A detector providing 2-level DOI information with equal crystal thickness also does not show significant gains. Finally, a 15 mm thick crystal leads to lower lesion detectability than a 20 mm thick crystal when keeping all other detector parameters (crystal width, timing resolution, and DOI capability) the same. However, improved timing performance with 15

  10. Improvement of the spatial resolution of the MicroPET R4 scanner by wobbling the bed.

    PubMed

    Suk, Joon Young; Thompson, Christopher J; Labuda, Aleks; Goertzen, Andrew L

    2008-04-01

    The MicroPET R4 scanner was designed for imaging small rodents such as mice and rats. In many cases the spatial resolution of this system is not sufficient for resolving structures of interest. In order to improve the spatial resolution of the MicroPET R4 through improved spatial sampling, the authors have implemented a variable radius eccentric motion, commonly referred to as wobbling, which is applied to the animal bed during scanning. The wobble motion is incorporated into the sinograms using modified histogramming software, capable of reading the bed wobble position from the list-mode data. The histogramming software corrects the data for the dwell time, apparent crystal location, and crystal-pair efficiency and applies a resolution matching filter. The data acquisition, reconstruction, and image display programs provided from the manufacturer required no modifications. For all studies a wobble period of 8 s was used. The optimal wobble radius was found to be 1.50 mm. The wobbled bed acquisition technique was tested by scanning a resolution phantom and a rat. Images from both studies acquired when using the wobble motion showed an improved spatial resolution when compared with comparable images acquired without the wobble motion. The bed wobbling mechanism can be added to any MicroPET system without major changes and without compromising any imaging modes. Implementing the wobble mechanism may represent a cost-effective method to upgrade the spatial resolution of a MicroPET when compared to the purchase of a newer generation system. PMID:18491514

  11. Characterization of a high-resolution hybrid DOI detector for a dedicated breast PET/CT scanner.

    PubMed

    Godinez, Felipe; Chaudhari, Abhijit J; Yang, Yongfeng; Farrell, Richard; Badawi, Ramsey D

    2012-06-01

    The aim of this study is to design and test a new high-resolution hybrid depth of interaction (DOI) detector for a dedicated breast PET/CT scanner. Two detectors have been designed and built. The completed detectors are based on a 14 × 14 array of 1.5 × 1.5 × 20 mm(3) unpolished lutetium orthosilicate scintillation crystals, with each element coated in a 50 μm layer of reflective material. The detector is read out from both ends using a position-sensitive photomultiplier tube (PSPMT) and a large active area (20 × 20 mm(2)) avalanche photodiode (APD) to enable acquisition of DOI information. Nuclear instrumentation modules were used to characterize the detectors' performances in terms of timing, intrinsic spatial resolution (ISR) and energy resolution, as well as DOI resolution with a dual-ended readout configuration. Measurements with the APD were performed at a temperature of 10 °C. All crystals were identified at all depths, even though the signal amplitude from the PSPMT decreases with depth away from it. We measured a timing resolution of 2.4 ns, and an average energy resolution of 19%. The mean ISR was measured to be 1.2 mm for crystals in the central row of the array for detectors in the face-to-face position. Two off-center positions were measured corresponding to 26° and 51° oblique photon incidence, and the mean ISR at these positions was 1.5 and 1.7 mm, respectively. The average DOI resolution across all crystals and depths was measured to be 2.9 mm (including the beam width of 0.6 mm). This detector design shows good promise as a high-resolution detector for a dedicated breast PET/CT scanner. PMID:22581109

  12. Use of [18F]fluorodeoxyglucose and the ATLAS small animal PET scanner to examine cerebral functional activation by whisker stimulation in unanesthetized rats

    PubMed Central

    Ravasi, Laura; Shimoji, Kazuaki; Soto-Montenegro, Marisa L.; Esaki, Takanori; Seidel, Jurgen; Sokoloff, Louis; Schmidt, Kathleen

    2011-01-01

    Purpose Stroking the whiskers of a rat is known to increase cerebral blood flow and glucose utilization in the somatosensory cortex. We sought to determine if this activation could be detected with small animal PET and [18F]FDG. Methods Awake rats were co-injected with [18F]FDG and [14C]DG, and during uptake of the tracers, either 5, 10, or 15 whiskers on one side of the face were continuously stimulated. At the end of uptake the animal was killed and imaged with the ATLAS small animal PET scanner. 14C autoradiography was then performed on brain sections obtained from each animal, and increases in tracer uptake in the somatosensory cortex were compared to those determined with PET. Results Both methods showed increases in [18F]FDG and [14C]DG uptake in the somatosensory cortex in response to stimulation of as few as five whiskers. However, the magnitude of activation determined from the PET images was less than that from autoradiography due to the lower spatial resolution of the PET scanner. Conclusions ATLAS small animal PET imaging with [18F]FDG can be used to assess neuronal functional activity in vivo. PMID:21326123

  13. Characterization of disease-related covariance topographies with SSMPCA toolbox: effects of spatial normalization and PET scanners.

    PubMed

    Peng, Shichun; Ma, Yilong; Spetsieris, Phoebe G; Mattis, Paul; Feigin, Andrew; Dhawan, Vijay; Eidelberg, David

    2014-05-01

    To generate imaging biomarkers from disease-specific brain networks, we have implemented a general toolbox to rapidly perform scaled subprofile modeling (SSM) based on principal component analysis (PCA) on brain images of patients and normals. This SSMPCA toolbox can define spatial covariance patterns whose expression in individual subjects can discriminate patients from controls or predict behavioral measures. The technique may depend on differences in spatial normalization algorithms and brain imaging systems. We have evaluated the reproducibility of characteristic metabolic patterns generated by SSMPCA in patients with Parkinson's disease (PD). We used [(18) F]fluorodeoxyglucose PET scans from patients with PD and normal controls. Motor-related (PDRP) and cognition-related (PDCP) metabolic patterns were derived from images spatially normalized using four versions of SPM software (spm99, spm2, spm5, and spm8). Differences between these patterns and subject scores were compared across multiple independent groups of patients and control subjects. These patterns and subject scores were highly reproducible with different normalization programs in terms of disease discrimination and cognitive correlation. Subject scores were also comparable in patients with PD imaged across multiple PET scanners. Our findings confirm a very high degree of consistency among brain networks and their clinical correlates in PD using images normalized in four different SPM platforms. SSMPCA toolbox can be used reliably for generating disease-specific imaging biomarkers despite the continued evolution of image preprocessing software in the neuroimaging community. Network expressions can be quantified in individual patients independent of different physical characteristics of PET cameras. PMID:23671030

  14. Comparison of Monte Carlo simulated and measured performance parameters of miniPET scanner

    NASA Astrophysics Data System (ADS)

    Kis, S. A.; Emri, M.; Opposits, G.; Bükki, T.; Valastyán, I.; Hegyesi, Gy.; Imrek, J.; Kalinka, G.; Molnár, J.; Novák, D.; Végh, J.; Kerek, A.; Trón, L.; Balkay, L.

    2007-02-01

    In vivo imaging of small laboratory animals is a valuable tool in the development of new drugs. For this purpose, miniPET, an easy to scale modular small animal PET camera has been developed at our institutes. The system has four modules, which makes it possible to rotate the whole detector system around the axis of the field of view. Data collection and image reconstruction are performed using a data acquisition (DAQ) module with Ethernet communication facility and a computer cluster of commercial PCs. Performance tests were carried out to determine system parameters, such as energy resolution, sensitivity and noise equivalent count rate. A modified GEANT4-based GATE Monte Carlo software package was used to simulate PET data analogous to those of the performance measurements. GATE was run on a Linux cluster of 10 processors (64 bit, Xeon with 3.0 GHz) and controlled by a SUN grid engine. The application of this special computer cluster reduced the time necessary for the simulations by an order of magnitude. The simulated energy spectra, maximum rate of true coincidences and sensitivity of the camera were in good agreement with the measured parameters.

  15. Validation of a Monte Carlo simulation of the Inveon PET scanner using GATE

    NASA Astrophysics Data System (ADS)

    Lu, Lijun; Zhang, Houjin; Bian, Zhaoying; Ma, Jianhua; Feng, Qiangjin; Chen, Wufan

    2016-08-01

    The purpose of this study is to validate the application of GATE (Geant4 Application for Tomographic Emission) Monte Carlo simulation toolkit in order to model the performance characteristics of Siemens Inveon small animal PET system. The simulation results were validated against experimental/published data in accordance with the NEMA NU-4 2008 protocol for standardized evaluation of spatial resolution, sensitivity, scatter fraction (SF) and noise equivalent counting rate (NECR) of a preclinical PET system. An agreement of less than 18% was obtained between the radial, tangential and axial spatial resolutions of the simulated and experimental results. The simulated peak NECR of mouse-size phantom agreed with the experimental result, while for the rat-size phantom simulated value was higher than experimental result. The simulated and experimental SFs of mouse- and rat- size phantom both reached an agreement of less than 2%. It has been shown the feasibility of our GATE model to accurately simulate, within certain limits, all major performance characteristics of Inveon PET system.

  16. Validity of using a 3-dimensional PET scanner during inhalation of 15O-labeled oxygen for quantitative assessment of regional metabolic rate of oxygen in man

    NASA Astrophysics Data System (ADS)

    Hori, Yuki; Hirano, Yoshiyuki; Koshino, Kazuhiro; Moriguchi, Tetsuaki; Iguchi, Satoshi; Yamamoto, Akihide; Enmi, Junichiro; Kawashima, Hidekazu; Zeniya, Tsutomu; Morita, Naomi; Nakagawara, Jyoji; Casey, Michael E.; Iida, Hidehiro

    2014-09-01

    Use of 15O labeled oxygen (15O2) and positron emission tomography (PET) allows quantitative assessment of the regional metabolic rate of oxygen (CMRO2) in vivo, which is essential to understanding the pathological status of patients with cerebral vascular and neurological disorders. The method has, however, been challenging, when a 3D PET scanner is employed, largely attributed to the presence of gaseous radioactivity in the trachea and the inhalation system, which results in a large amount of scatter and random events in the PET assessment. The present study was intended to evaluate the adequacy of using a recently available commercial 3D PET scanner in the assessment of regional cerebral radioactivity distribution during an inhalation of 15O2. Systematic experiments were carried out on a brain phantom. Experiments were also performed on a healthy volunteer following a recently developed protocol for simultaneous assessment of CMRO2 and cerebral blood flow, which involves sequential administration of 15O2 and C15O2. A particular intention was to evaluate the adequacy of the scatter-correction procedures. The phantom experiment demonstrated that errors were within 3% at the practically maximum radioactivity in the face mask, with the greatest radioactivity in the lung. The volunteer experiment demonstrated that the counting rate was at peak during the 15O gas inhalation period, within a verified range. Tomographic images represented good quality over the entire FOV, including the lower part of the cerebral structures and the carotid artery regions. The scatter-correction procedures appeared to be important, particularly in the process to compensate for the scatter originating outside the FOV. Reconstructed images dramatically changed if the correction was carried out using inappropriate procedures. This study demonstrated that accurate reconstruction could be obtained when the scatter compensation was appropriately carried out. This study also suggested the

  17. Investigation of time-of-flight benefits in an LYSO-based PET/CT scanner: A Monte Carlo study using GATE

    NASA Astrophysics Data System (ADS)

    Geramifar, P.; Ay, M. R.; Shamsaie Zafarghandi, M.; Sarkar, S.; Loudos, G.; Rahmim, A.

    2011-06-01

    The advent of fast scintillators yielding great light yield and/or stopping power, along with advances in photomultiplier tubes and electronics, have rekindled interest in time-of-flight (TOF) PET. Because the potential performance improvements offered by TOF PET are substantial, efforts to improve PET timing should prove very fruitful. In this study, we performed Monte Carlo simulations to explore what gains in PET performance could be achieved if the coincidence resolving time (CRT) in the LYSO-based PET component of Discovery RX PET/CT scanner were improved. For this purpose, the GATE Monte Carlo package was utilized, providing the ability to model and characterize various physical phenomena in PET imaging. For the present investigation, count rate performance and signal to noise ratio (SNR) values in different activity concentrations were simulated for different coincidence timing windows of 4, 5.85, 6, 6.5, 8, 10 and 12 ns and with different CRTs of 100-900 ps FWHM involving 50 ps FWHM increments using the NEMA scatter phantom. Strong evidence supporting robustness of the simulations was found as observed in the good agreement between measured and simulated data for the cases of estimating axial sensitivity, axial and transaxial detection position, gamma non-collinearity angle distribution and positron annihilation distance. In the non-TOF context, the results show that the random event rate can be reduced by using narrower coincidence timing window widths, demonstrating considerable enhancements in the peak noise equivalent count rate (NECR) performance. The peak NECR had increased by ˜50% when utilizing the coincidence window width of 4 ns. At the same time, utilization of TOF information resulted in improved NECR and SNR with the dramatic reduction of random coincidences as a function of CRT. For example, with CRT of 500 ps FWHM, a factor of 2.3 reduction in random rates, factor of 1.5 increase in NECR and factor of 2.1 improvement in SNR is achievable

  18. Simultaneous hyperpolarized 13C-pyruvate MRI and 18F-FDG-PET in cancer (hyperPET): feasibility of a new imaging concept using a clinical PET/MRI scanner

    PubMed Central

    Gutte, Henrik; Hansen, Adam E; Henriksen, Sarah T; Johannesen, Helle H; Ardenkjaer-Larsen, Jan; Vignaud, Alexandre; Hansen, Anders E; Børresen, Betina; Klausen, Thomas L; Wittekind, Anne-Mette N; Gillings, Nic; Kristensen, Annemarie T; Clemmensen, Andreas; Højgaard, Liselotte; Kjær, Andreas

    2015-01-01

    In this paper we demonstrate, for the first time, the feasibility of a new imaging concept - combined hyperpolarized 13C-pyruvate magnetic resonance spectroscopic imaging (MRSI) and 18F-FDG-PET imaging. This procedure was performed in a clinical PET/MRI scanner with a canine cancer patient. We have named this concept hyper PET. Intravenous injection of the hyperpolarized 13C-pyruvate results in an increase of 13C-lactate, 13C-alanine and 13C-CO2 (13C-HCO3) resonance peaks relative to the tissue, disease and the metabolic state probed. Accordingly, with dynamic nuclear polarization (DNP) and use of 13C-pyruvate it is now possible to directly study the Warburg Effect through the rate of conversion of 13C-pyruvate to 13C-lactate. In this study, we combined it with 18F-FDG-PET that studies uptake of glucose in the cells. A canine cancer patient with a histology verified local recurrence of a liposarcoma on the right forepaw was imaged using a combined PET/MR clinical scanner. PET was performed as a single-bed, 10 min acquisition, 107 min post injection of 310 MBq 18F-FDG. 13C-chemical shift imaging (CSI) was performed just after FDG-PET and 30 s post injection of 23 mL hyperpolarized 13C-pyruvate. Peak heights of 13C-pyruvate and 13C-lactate were quantified using a general linear model. Anatomic 1H-MRI included axial and coronal T1 vibe, coronal T2-tse and axial T1-tse with fat saturation following gadolinium injection. In the tumor we found clearly increased 13C-lactate production, which also corresponded to high 18F-FDG uptake on PET. This is in agreement with the fact that glycolysis and production of lactate are increased in tumor cells compared to normal cells. Yet, most interestingly, also in the muscle of the forepaw of the dog high 18F-FDG uptake was observed. This was due to activity in these muscles prior to anesthesia, which was not accompanied by a similarly high 13C-lactate production. Accordingly, this clearly demonstrates how the Warburg Effect directly

  19. Clinical count rate performance of an LSO PET/CT scanner utilizing a new front-end electronics architecture with sub-nanosecond intrinsic timing resolution

    NASA Astrophysics Data System (ADS)

    Carney, J. P. J.; Townsend, D. W.

    2006-12-01

    A new front-end electronics architecture with sub-nanosecond intrinsic timing resolution has recently been incorporated into a 16 slice LSO PET/CT scanner for imaging applications in oncology. The new electronics are designed to work optimally with the lutetium orthosilicate (LSO) scintillator. Clinical performance of the LSO PET/CT is examined before and after upgrading to the new PICO 3D electronics, and compared with results using the NEMA NU 2 standard for evaluating scanner performance. Improved noise-equivalent count rates are seen in clinical studies, and reduced scatter fractions are observed, consistent with the increased lower-level energy threshold used to reject scatter events in the upgraded configuration.

  20. The countrate performance of the volume imaging PENN-PET scanner

    SciTech Connect

    Smith, R.J.; Karp, J.S. . Dept. of Radiology); Muehllehner, G. )

    1994-12-01

    The UGM PENN-PET Camera uses large position sensitive detectors and operates without septa. This design results in high sensitivity and 3-D imaging capability, but poses problems in high countrate situations. The maximum true countrates and random countrates have been measured, as a function of object size in the field-of-view. The countrate performance is understood in terms of the limiting process rates and event rejection in the camera. In addition, the camera is calibrated to generate absolute activity concentrations to within 5% by correcting for system deadtime with up to 3 mCi in the field of view in brain studies. This allows the performance of a variety of brain and body studies, and accurate quantitation of the data over a wide-range of imaging countrates, from [sup 18]F-FDG brain studies to [sup 15]O-water dynamic brain studies.

  1. CT with a CMOS flat panel detector integrated on the YAP-(S)PET scanner for in vivo small animal imaging

    NASA Astrophysics Data System (ADS)

    Di Domenico, Giovanni; Cesca, Nicola; Zavattini, Guido; Auricchio, Natalia; Gambaccini, Mauro

    2007-02-01

    Several research groups are pursuing multimodality simultaneous functional and morphological imaging. In this line of research the high resolution YAP-(S)PET small animal integrated PET-SPECT imaging system, constructed by our group of medical physics at the University of Ferrara, is being upgraded with a computed tomography (CT). In this way it will be possible to perform in vivo molecular and genomic imaging studies on small animals (such as mice and rats) and at the same time obtain morphological information necessary for both attenuation correction and accurate localization of the region under investigation. We have take simultaneous PET-CT and SPECT-CT images of phantoms obtained with a single scanner.

  2. A rat head holder for simultaneous scanning of two rats in small animal PET scanners: Design, construction, feasibility testing and kinetic validation

    PubMed Central

    Cheng, Tee Ean; Yoder, Karmen K.; Normandin, Marc D.; Risacher, Shannon L.; Converse, Alexander K.; Hampel, Joseph A.; Miller, Michael A.; Morris, Evan D.

    2013-01-01

    To reduce imaging costs, we designed a head holder for scanning two rats simultaneously in small animal PET scanners. Our goals were (i) to maintain high sensitivity and (ii) to minimize repositioning error between scans. Methods A semi-stereotaxic dual rat head holder was designed and constructed for dual rat scanning in our IndyPET-II scanner and the commercial microPET P4. It was also used for single rat scanning in a small-bore, high-resolution animal scanner (“ISAP”). Positional repeatability was validated via multiple [11C]Raclopride scans of a single rat on different days. Accuracy of repositioning was determined by visual comparison of images, and by metrics derived through image alignment. Kinetic validation was assessed via analysis of [18F]Fluorodeoxyglucose ([18F]FDG) dynamic PET studies of six rats. Each rat was scanned twice: once individually, with brain positioned at the center of field of view (CFOV), and once with a partner, with brain away from CFOV. Both rats were injected with FDG during each dual rat session. Patlak uptake constants (Ki) were calculated from whole brain images. Effects of attenuation and scatter correction on single versus dual scan images were explored. Results Image comparison and alignment metrics indicated excellent repositioning of rats. Scaled time-activity-curves from single and dual rat scans were indistinguishable. Average single and dual scan Ki values differed by only 6.3 ± 7.5%. Conclusion Dual rat scanning in a semi-sterotaxic holder is practical for economical small animal scanning and does not compromise kinetic accuracy of [18F]FDG dynamic scan data. PMID:18824025

  3. Experimental evaluation and basis function optimization of the spatially variant image-space PSF on the Ingenuity PET/MR scanner

    SciTech Connect

    Kotasidis, Fotis A.; Zaidi, Habib

    2014-06-15

    Purpose: The Ingenuity time-of-flight (TF) PET/MR is a recently developed hybrid scanner combining the molecular imaging capabilities of PET with the excellent soft tissue contrast of MRI. It is becoming common practice to characterize the system's point spread function (PSF) and understand its variation under spatial transformations to guide clinical studies and potentially use it within resolution recovery image reconstruction algorithms. Furthermore, due to the system's utilization of overlapping and spherical symmetric Kaiser-Bessel basis functions during image reconstruction, its image space PSF and reconstructed spatial resolution could be affected by the selection of the basis function parameters. Hence, a detailed investigation into the multidimensional basis function parameter space is needed to evaluate the impact of these parameters on spatial resolution. Methods: Using an array of 12 × 7 printed point sources, along with a custom made phantom, and with the MR magnet on, the system's spatially variant image-based PSF was characterized in detail. Moreover, basis function parameters were systematically varied during reconstruction (list-mode TF OSEM) to evaluate their impact on the reconstructed resolution and the image space PSF. Following the spatial resolution optimization, phantom, and clinical studies were subsequently reconstructed using representative basis function parameters. Results: Based on the analysis and under standard basis function parameters, the axial and tangential components of the PSF were found to be almost invariant under spatial transformations (∼4 mm) while the radial component varied modestly from 4 to 6.7 mm. Using a systematic investigation into the basis function parameter space, the spatial resolution was found to degrade for basis functions with a large radius and small shape parameter. However, it was found that optimizing the spatial resolution in the reconstructed PET images, while having a good basis function

  4. SU-D-9A-04: Brain PET/CT Imaging On a Scanner with a Large Axial Field-Of-View

    SciTech Connect

    Park, M; Gerbaudo, V; Hamberg, L; Seaver, K; Kijewski, M

    2014-06-01

    Purpose: Large axial field-of-view (FOV) PET/CT scanners are valued for high sensitivity. Brain PET image quality may depend on the head position within the FOV. We investigated the precision of activity estimation for brain PET imaging when the brain was positioned at the end (END) and in the middle (CEN) of the FOV. The additional CT dose for the CEN position was recorded. Methods: An image quality (Jaszczak) phantom and a striatal phantom were filled with F-18 and positioned in END and CEN locations. For each phantom and each location, we acquired a ∼1-hr listmode PET, rebinned the data into 10 frames with equal number of coincidence events, and reconstructed each frame using an iterative algorithm. For the striatal phantom, END and CEN were compared by drawing on each image three regions of interest (ROI) in axially separated uniform areas. The standard deviation of the activity estimation within each ROI was averaged over the 10 images. The coefficient of variation (CV) for activity estimation was calculated at each position. Image quality was assessed by inspecting the resolution bar pattern in the Jaszczak phantom at two different head positions. Results: The CV was the lowest for ROIs near the center of the FOV. For slices near the end, not only was the CV highest, but also the resolution pattern was degraded. CTDIvol summarized in the dose report indicated that the CT dose was ∼ 10% higher for CEN as compared to END position. Conclusion: Positioning the brain in the middle of the FOV in a large FOV PET/CT scanner allows more precise measurement of tracer uptake and better image quality at the cost of increased CT dose. For the end location longer scan times may minimize image quality degradation without any additional CT dose.

  5. Cold wall effect eliminating method to determine the contrast recovery coefficient for small animal PET scanners using the NEMA NU-4 image quality phantom

    NASA Astrophysics Data System (ADS)

    Lajtos, Imre; Czernin, Johannes; Dahlbom, Magnus; Daver, Freddie; Emri, Miklos; Farshchi-Heydari, Salman; Forgacs, Attila; Hoh, Carl K.; Joszai, Istvan; Krizsan, Aron K.; Lantos, Judit; Major, Peter; Molnar, Jozsef; Opposits, Gabor; Tron, Lajos; Vera, David R.; Balkay, Laszlo

    2014-06-01

    The contrast recovery coefficients (CRC) were evaluated for five different small animal PET scanners: GE Explore Vista, Genisys4, MiniPET-2, nanoScan PC and Siemens Inveon. The NEMA NU-4 2008 performance test with the suggested image quality phantom (NU4IQ) does not allow the determination of the CRC values for the hot regions in the phantom. This drawback of NU4IQ phantom motivated us to develop a new method for this purpose. The method includes special acquisition and reconstruction protocols using the original phantom, and results in an artificially merged image enabling the evaluation of CRC values. An advantageous feature of this method is that it stops the cold wall effect from distorting the CRC calculation. Our suggested protocol results in a set of CRC values contributing to the characterization of small animal PET scanners. GATE simulations were also performed to validate the new method and verify the evaluated CRC values. We also demonstrated that the numerical values of this parameter depend on the actual object contrast of the hot region(s) and this mainly comes from the spillover effect. This effect was also studied while analysing the background activity level around the hot rods. We revealed that the calculated background mean values depended on the target contrast in a scanner specific manner. Performing the artificially merged imaging procedure and additional simulations using the micro hollow sphere (MHS) phantom geometry, we also proved that the inactive wall around the hot spheres can have a remarkable impact on the calculated CRC. In conclusion, we have shown that the proposed artificial merging procedure and the commonly used NU4IQ phantom prescribed by the NEMA NU-4 can easily deliver reliable CRC data otherwise unavailable for the NU4IQ phantom in the conventional protocol or the MHS phantom.

  6. In vivo quantitative imaging of photoassimilate transport dynamics and allocation in large plants using a commercial positron emission tomography (PET) scanner

    SciTech Connect

    Karve, Abhijit A.; Alexoff, David; Kim, Dohyun; Schueller, Michael J.; Ferrieri, Richard A.; Babst, Benjamin A.

    2015-11-09

    Although important aspects of whole-plant carbon allocation in crop plants (e.g., to grain) occur late in development when the plants are large, techniques to study carbon transport and allocation processes have not been adapted for large plants. Positron emission tomography (PET), developed for dynamic imaging in medicine, has been applied in plant studies to measure the transport and allocation patterns of carbohydrates, nutrients, and phytohormones labeled with positron-emitting radioisotopes. However, the cost of PET and its limitation to smaller plants has restricted its use in plant biology. Here we describe the adaptation and optimization of a commercial clinical PET scanner to measure transport dynamics and allocation patterns of 11C-photoassimilates in large crops. Based on measurements of a phantom, we optimized instrument settings, including use of 3-D mode and attenuation correction to maximize the accuracy of measurements. To demonstrate the utility of PET, we measured 11C-photoassimilate transport and allocation in Sorghum bicolor, an important staple crop, at vegetative and reproductive stages (40 and 70 days after planting; DAP). The 11C-photoassimilate transport speed did not change over the two developmental stages. However, within a stem, transport speeds were reduced across nodes, likely due to higher 11C-photoassimilate unloading in the nodes. Photosynthesis in leaves and the amount of 11C that was exported to the rest of the plant decreased as plants matured. In young plants, exported 11C was allocated mostly (88 %) to the roots and stem, but in flowering plants (70 DAP) the majority of the exported 11C (64 %) was allocated to the apex. Our results show that commercial PET scanners can be used reliably to measure whole-plant C-allocation in large plants nondestructively including, importantly, allocation to roots in soil. This capability revealed extreme changes in

  7. Simultaneous scanning of two mice in a small-animal PET scanner: a simulation-based assessment of the signal degradation.

    PubMed

    Reilhac, Anthonin; Boisson, Frédéric; Wimberley, Catriona; Parmar, Arvind; Zahra, David; Hamze, Hasar; Davis, Emma; Arthur, Andrew; Bouillot, Caroline; Charil, Arnaud; Grégoire, Marie-Claude

    2016-02-01

    In PET imaging, research groups have recently proposed different experimental set ups allowing multiple animals to be simultaneously imaged in a scanner in order to reduce the costs and increase the throughput. In those studies, the technical feasibility was demonstrated and the signal degradation caused by additional mice in the FOV characterized, however, the impact of the signal degradation on the outcome of a PET study has not yet been studied. Here we thoroughly investigated, using Monte Carlo simulated [18F]FDG and [11C]Raclopride PET studies, different experimental designs for whole-body and brain acquisitions of two mice and assessed the actual impact on the detection of biological variations as compared to a single-mouse setting. First, we extended the validation of the PET-SORTEO Monte Carlo simulation platform for the simultaneous simulation of two animals. Then, we designed [18F]FDG and [11C]Raclopride input mouse models for the simulation of realistic whole-body and brain PET studies. Simulated studies allowed us to accurately estimate the differences in detection between single- and dual-mode acquisition settings that are purely the result of having two animals in the FOV. Validation results showed that PET-SORTEO accurately reproduced the spatial resolution and noise degradations that were observed with actual dual phantom experiments. The simulated [18F]FDG whole-body study showed that the resolution loss due to the off-center positioning of the mice was the biggest contributing factor in signal degradation at the pixel level and a minimal inter-animal distance as well as the use of reconstruction methods with resolution modeling should be preferred. Dual mode acquisition did not have a major impact on ROI-based analysis except in situations where uptake values in organs from the same subject were compared. The simulated [11C]Raclopride study however showed that dual-mice imaging strongly reduced the sensitivity to variations when mice were

  8. Simultaneous scanning of two mice in a small-animal PET scanner: a simulation-based assessment of the signal degradation

    NASA Astrophysics Data System (ADS)

    Reilhac, Anthonin; Boisson, Frédéric; Wimberley, Catriona; Parmar, Arvind; Zahra, David; Hamze, Hasar; Davis, Emma; Arthur, Andrew; Bouillot, Caroline; Charil, Arnaud; Grégoire, Marie-Claude

    2016-02-01

    In PET imaging, research groups have recently proposed different experimental set ups allowing multiple animals to be simultaneously imaged in a scanner in order to reduce the costs and increase the throughput. In those studies, the technical feasibility was demonstrated and the signal degradation caused by additional mice in the FOV characterized, however, the impact of the signal degradation on the outcome of a PET study has not yet been studied. Here we thoroughly investigated, using Monte Carlo simulated [18F]FDG and [11C]Raclopride PET studies, different experimental designs for whole-body and brain acquisitions of two mice and assessed the actual impact on the detection of biological variations as compared to a single-mouse setting. First, we extended the validation of the PET-SORTEO Monte Carlo simulation platform for the simultaneous simulation of two animals. Then, we designed [18F]FDG and [11C]Raclopride input mouse models for the simulation of realistic whole-body and brain PET studies. Simulated studies allowed us to accurately estimate the differences in detection between single- and dual-mode acquisition settings that are purely the result of having two animals in the FOV. Validation results showed that PET-SORTEO accurately reproduced the spatial resolution and noise degradations that were observed with actual dual phantom experiments. The simulated [18F]FDG whole-body study showed that the resolution loss due to the off-center positioning of the mice was the biggest contributing factor in signal degradation at the pixel level and a minimal inter-animal distance as well as the use of reconstruction methods with resolution modeling should be preferred. Dual mode acquisition did not have a major impact on ROI-based analysis except in situations where uptake values in organs from the same subject were compared. The simulated [11C]Raclopride study however showed that dual-mice imaging strongly reduced the sensitivity to variations when mice were

  9. Toward VIP-PIX: A Low Noise Readout ASIC for Pixelated CdTe Gamma-Ray Detectors for Use in the Next Generation of PET Scanners.

    PubMed

    Macias-Montero, Jose-Gabriel; Sarraj, Maher; Chmeissani, Mokhtar; Puigdengoles, Carles; Lorenzo, Gianluca De; Martínez, Ricardo

    2013-08-01

    VIP-PIX will be a low noise and low power pixel readout electronics with digital output for pixelated Cadmium Telluride (CdTe) detectors. The proposed pixel will be part of a 2D pixel-array detector for various types of nuclear medicine imaging devices such as positron-emission tomography (PET) scanners, Compton gamma cameras, and positron-emission mammography (PEM) scanners. Each pixel will include a SAR ADC that provides the energy deposited with 10-bit resolution. Simultaneously, the self-triggered pixel which will be connected to a global time-to-digital converter (TDC) with 1 ns resolution will provide the event's time stamp. The analog part of the readout chain and the ADC have been fabricated with TSMC 0.25 μm mixed-signal CMOS technology and characterized with an external test pulse. The power consumption of these parts is 200 μW from a 2.5 V supply. It offers 4 switchable gains from ±10 mV/fC to ±40 mV/fC and an input charge dynamic range of up to ±70 fC for the minimum gain for both polarities. Based on noise measurements, the expected equivalent noise charge (ENC) is 65 e(-) RMS at room temperature. PMID:24187382

  10. Toward VIP-PIX: A Low Noise Readout ASIC for Pixelated CdTe Gamma-Ray Detectors for Use in the Next Generation of PET Scanners

    PubMed Central

    Macias-Montero, Jose-Gabriel; Sarraj, Maher; Chmeissani, Mokhtar; Puigdengoles, Carles; Lorenzo, Gianluca De; Martínez, Ricardo

    2013-01-01

    VIP-PIX will be a low noise and low power pixel readout electronics with digital output for pixelated Cadmium Telluride (CdTe) detectors. The proposed pixel will be part of a 2D pixel-array detector for various types of nuclear medicine imaging devices such as positron-emission tomography (PET) scanners, Compton gamma cameras, and positron-emission mammography (PEM) scanners. Each pixel will include a SAR ADC that provides the energy deposited with 10-bit resolution. Simultaneously, the self-triggered pixel which will be connected to a global time-to-digital converter (TDC) with 1 ns resolution will provide the event’s time stamp. The analog part of the readout chain and the ADC have been fabricated with TSMC 0.25 μm mixed-signal CMOS technology and characterized with an external test pulse. The power consumption of these parts is 200 μW from a 2.5 V supply. It offers 4 switchable gains from ±10 mV/fC to ±40 mV/fC and an input charge dynamic range of up to ±70 fC for the minimum gain for both polarities. Based on noise measurements, the expected equivalent noise charge (ENC) is 65 e− RMS at room temperature. PMID:24187382

  11. OSSI-PET: Open-Access Database of Simulated [(11)C]Raclopride Scans for the Inveon Preclinical PET Scanner: Application to the Optimization of Reconstruction Methods for Dynamic Studies.

    PubMed

    Garcia, Marie-Paule; Charil, Arnaud; Callaghan, Paul; Wimberley, Catriona; Busso, Florian; Gregoire, Marie-Claude; Bardies, Manuel; Reilhac, Anthonin

    2016-07-01

    A wide range of medical imaging applications benefits from the availability of realistic ground truth data. In the case of positron emission tomography (PET), ground truth data is crucial to validate processing algorithms and assessing their performances. The design of such ground truth data often relies on Monte-Carlo simulation techniques. Since the creation of a large dataset is not trivial both in terms of computing time and realism, we propose the OSSI-PET database containing 350 simulated [(11)C]Raclopride dynamic scans for rats, created specifically for the Inveon pre-clinical PET scanner. The originality of this database lies on the availability of several groups of scans with controlled biological variations in the striata. Besides, each group consists of a large number of realizations (i.e., noise replicates). We present the construction methodology of this database using rat pharmacokinetic and anatomical models. A first application using the OSSI-PET database is presented. Several commonly used reconstruction techniques were compared in terms of image quality, accuracy and variability of the activity estimates and of the computed kinetic parameters. The results showed that OP-OSEM3D iterative reconstruction method outperformed the other tested methods. Analytical methods such as FBP2D and 3DRP also produced satisfactory results. However, FORE followed by OSEM2D reconstructions should be avoided. Beyond the illustration of the potential of the database, this application will help scientists to understand the different sources of noise and bias that can occur at the different steps in the processing and will be very useful for choosing appropriate reconstruction methods and parameters. PMID:26863655

  12. Development of a prototype PET scanner with depth-of-interaction measurement using solid-state photomultiplier arrays and parallel readout electronics

    PubMed Central

    Shao, Yiping; Sun, Xishan; Lan, Kejian A.; Bircher, Chad; Lou, Kai; Deng, Zhi

    2014-01-01

    DOI-measureable PET scanner. PMID:24556629

  13. Development of a prototype PET scanner with depth-of-interaction measurement using solid-state photomultiplier arrays and parallel readout electronics.

    PubMed

    Shao, Yiping; Sun, Xishan; Lan, Kejian A; Bircher, Chad; Lou, Kai; Deng, Zhi

    2014-03-01

    -measureable PET scanner. PMID:24556629

  14. Development of a prototype PET scanner with depth-of-interaction measurement using solid-state photomultiplier arrays and parallel readout electronics

    NASA Astrophysics Data System (ADS)

    Shao, Yiping; Sun, Xishan; Lan, Kejian A.; Bircher, Chad; Lou, Kai; Deng, Zhi

    2014-03-01

    -measureable PET scanner.

  15. In vivo quantitative imaging of photoassimilate transport dynamics and allocation in large plants using a commercial positron emission tomography (PET) scanner

    DOE PAGESBeta

    Karve, Abhijit A.; Alexoff, David; Kim, Dohyun; Schueller, Michael J.; Ferrieri, Richard A.; Babst, Benjamin A.

    2015-11-09

    Although important aspects of whole-plant carbon allocation in crop plants (e.g., to grain) occur late in development when the plants are large, techniques to study carbon transport and allocation processes have not been adapted for large plants. Positron emission tomography (PET), developed for dynamic imaging in medicine, has been applied in plant studies to measure the transport and allocation patterns of carbohydrates, nutrients, and phytohormones labeled with positron-emitting radioisotopes. However, the cost of PET and its limitation to smaller plants has restricted its use in plant biology. Here we describe the adaptation and optimization of a commercial clinical PET scannermore » to measure transport dynamics and allocation patterns of 11C-photoassimilates in large crops. Based on measurements of a phantom, we optimized instrument settings, including use of 3-D mode and attenuation correction to maximize the accuracy of measurements. To demonstrate the utility of PET, we measured 11C-photoassimilate transport and allocation in Sorghum bicolor, an important staple crop, at vegetative and reproductive stages (40 and 70 days after planting; DAP). The 11C-photoassimilate transport speed did not change over the two developmental stages. However, within a stem, transport speeds were reduced across nodes, likely due to higher 11C-photoassimilate unloading in the nodes. Photosynthesis in leaves and the amount of 11C that was exported to the rest of the plant decreased as plants matured. In young plants, exported 11C was allocated mostly (88 %) to the roots and stem, but in flowering plants (70 DAP) the majority of the exported 11C (64 %) was allocated to the apex. Our results show that commercial PET scanners can be used reliably to measure whole-plant C-allocation in large plants nondestructively including, importantly, allocation to roots in soil. This capability revealed extreme changes in carbon allocation in sorghum plants, as they advanced to maturity

  16. The effect of activity outside the field of view on image quality for a 3D LSO-based whole body PET/CT scanner.

    PubMed

    Matheoud, R; Secco, C; Della Monica, P; Leva, L; Sacchetti, G; Inglese, E; Brambilla, M

    2009-10-01

    The purpose of this study was to quantify the influence of outside field of view (FOV) activity concentration (A(c)(,out)) on the noise equivalent count rate (NECR), scatter fraction (SF) and image quality of a 3D LSO whole-body PET/CT scanner. The contrast-to-noise ratio (CNR) was the figure of merit used to characterize the image quality of PET scans. A modified International Electrotechnical Commission (IEC) phantom was used to obtain SF and counting rates similar to those found in average patients. A scatter phantom was positioned at the end of the modified IEC phantom to simulate an activity that extends beyond the scanner. The modified IEC phantom was filled with (18)F (11 kBq mL(-1)) and the spherical targets, with internal diameter (ID) ranging from 10 to 37 mm, had a target-to-background ratio of 10. PET images were acquired with background activity concentrations into the FOV (A(c)(,bkg)) about 11, 9.2, 6.6, 5.2 and 3.5 kBq mL(-1). The emission scan duration (ESD) was set to 1, 2, 3 and 4 min. The tube inside the scatter phantom was filled with activities to provide A(c)(,out) in the whole scatter phantom of zero, half, unity, twofold and fourfold the one of the modified IEC phantom. Plots of CNR versus the various parameters are provided. Multiple linear regression was employed to study the effects of A(c)(,out) on CNR, adjusted for the presence of variables (sphere ID, A(c)(,bkg) and ESD) related to CNR. The presence of outside FOV activity at the same concentration as the one inside the FOV reduces peak NECR of 30%. The increase in SF is marginal (1.2%). CNR diminishes significantly with increasing outside FOV activity, in the range explored. ESD and A(c)(,out) have a similar weight in accounting for CNR variance. Thus, an experimental law that adjusts the scan duration to the outside FOV activity can be devised. Recovery of CNR loss due to an elevated A(c)(,out) activity seems feasible by modulating the ESD in individual bed positions according to A

  17. Single scan parameterization of space-variant point spread functions in image space via a printed array: the impact for two PET/CT scanners

    NASA Astrophysics Data System (ADS)

    Kotasidis, F. A.; Matthews, J. C.; Angelis, G. I.; Noonan, P. J.; Jackson, A.; Price, P.; Lionheart, W. R.; Reader, A. J.

    2011-05-01

    Incorporation of a resolution model during statistical image reconstruction often produces images of improved resolution and signal-to-noise ratio. A novel and practical methodology to rapidly and accurately determine the overall emission and detection blurring component of the system matrix using a printed point source array within a custom-made Perspex phantom is presented. The array was scanned at different positions and orientations within the field of view (FOV) to examine the feasibility of extrapolating the measured point source blurring to other locations in the FOV and the robustness of measurements from a single point source array scan. We measured the spatially-variant image-based blurring on two PET/CT scanners, the B-Hi-Rez and the TruePoint TrueV. These measured spatially-variant kernels and the spatially-invariant kernel at the FOV centre were then incorporated within an ordinary Poisson ordered subset expectation maximization (OP-OSEM) algorithm and compared to the manufacturer's implementation using projection space resolution modelling (RM). Comparisons were based on a point source array, the NEMA IEC image quality phantom, the Cologne resolution phantom and two clinical studies (carbon-11 labelled anti-sense oligonucleotide [11C]-ASO and fluorine-18 labelled fluoro-l-thymidine [18F]-FLT). Robust and accurate measurements of spatially-variant image blurring were successfully obtained from a single scan. Spatially-variant resolution modelling resulted in notable resolution improvements away from the centre of the FOV. Comparison between spatially-variant image-space methods and the projection-space approach (the first such report, using a range of studies) demonstrated very similar performance with our image-based implementation producing slightly better contrast recovery (CR) for the same level of image roughness (IR). These results demonstrate that image-based resolution modelling within reconstruction is a valid alternative to projection

  18. Scanner Art

    ERIC Educational Resources Information Center

    Jaworski, Joy; Murphy, Kris

    2009-01-01

    In this article, the authors describe how they incorporated environmental awareness into their art curriculum. Here, they describe a digital photography project in which their students used flatbed scanners as cameras. Their students composed their objects directly on the scanner. The lesson enabled students to realize that artists have voices…

  19. Cylindrical Scanner

    Energy Science and Technology Software Center (ESTSC)

    1999-04-29

    The CS system is designed to provide a very fast imaging system in order to search for weapons on persons in an airport environment. The Cylindrical Scanner moves a vertical transceiver array rapidly around a person standing stationary. The software can be segmented in to three specific tasks. The first task is data acquisition and scanner control. At the operator's request, this task commands the scanner to move and the radar transceiver array to sendmore » data to the computer system in a known and well-ordered manner. The array is moved over the complete aperture in 10 to 12 seconds. At the completion of the array movement the second software task automatically reconstructs the high-resolution image from the radar data utilizing the integrated DSP boards. The third task displays the resulting images, as they become available, to the computer screen for user review and analysis.« less

  20. Cylindrical Scanner

    SciTech Connect

    Hall, Thomas E.

    1999-04-29

    The CS system is designed to provide a very fast imaging system in order to search for weapons on persons in an airport environment. The Cylindrical Scanner moves a vertical transceiver array rapidly around a person standing stationary. The software can be segmented in to three specific tasks. The first task is data acquisition and scanner control. At the operator's request, this task commands the scanner to move and the radar transceiver array to send data to the computer system in a known and well-ordered manner. The array is moved over the complete aperture in 10 to 12 seconds. At the completion of the array movement the second software task automatically reconstructs the high-resolution image from the radar data utilizing the integrated DSP boards. The third task displays the resulting images, as they become available, to the computer screen for user review and analysis.

  1. Monte Carlo simulations in multi-detector CT (MDCT) for two PET/CT scanner models using MASH and FASH adult phantoms

    NASA Astrophysics Data System (ADS)

    Belinato, W.; Santos, W. S.; Paschoal, C. M. M.; Souza, D. N.

    2015-06-01

    The combination of positron emission tomography (PET) and computed tomography (CT) has been extensively used in oncology for diagnosis and staging of tumors, radiotherapy planning and follow-up of patients with cancer, as well as in cardiology and neurology. This study determines by the Monte Carlo method the internal organ dose deposition for computational phantoms created by multidetector CT (MDCT) beams of two PET/CT devices operating with different parameters. The different MDCT beam parameters were largely related to the total filtration that provides a beam energetic change inside the gantry. This parameter was determined experimentally with the Accu-Gold Radcal measurement system. The experimental values of the total filtration were included in the simulations of two MCNPX code scenarios. The absorbed organ doses obtained in MASH and FASH phantoms indicate that bowtie filter geometry and the energy of the X-ray beam have significant influence on the results, although this influence can be compensated by adjusting other variables such as the tube current-time product (mAs) and pitch during PET/CT procedures.

  2. Characterization and optimization of image quality as a function of reconstruction algorithms and parameter settings in a Siemens Inveon small-animal PET scanner using the NEMA NU 4-2008 standards

    NASA Astrophysics Data System (ADS)

    Visser, Eric P.; Disselhorst, Jonathan A.; van Lier, Monique G. J. T. B.; Laverman, Peter; de Jong, Gabie M.; Oyen, Wim J. G.; Boerman, Otto C.

    2011-02-01

    The image reconstruction algorithms provided with the Siemens Inveon small-animal PET scanner are filtered backprojection (FBP), 3-dimensional reprojection (3DRP), ordered subset expectation maximization in 2 or 3 dimensions (OSEM2D/3D) and maximum a posteriori (MAP) reconstruction. This study aimed at optimizing the reconstruction parameter settings with regard to image quality (IQ) as defined by the NEMA NU 4-2008 standards. The NEMA NU 4-2008 image quality phantom was used to determine image noise, expressed as percentage standard deviation in the uniform phantom region (%STD unif), activity recovery coefficients for the FDG-filled rods (RC rod), and spill-over ratios for the non-radioactive water- and air-filled phantom compartments (SOR wat and SOR air). Although not required by NEMA NU 4, we also determined a contrast-to-noise ratio for each rod (CNR rod), expressing the trade-off between activity recovery and image noise. For FBP and 3DRP the cut-off frequency of the applied filters, and for OSEM2D and OSEM3D, the number of iterations was varied. For MAP, the "smoothing parameter" β and the type of uniformity constraint (variance or resolution) were varied. Results of these analyses were demonstrated in images of an FDG-injected rat showing tumours in the liver, and of a mouse injected with an 18F-labeled peptide, showing a small subcutaneous tumour and the cortex structure of the kidneys. Optimum IQ in terms of CNR rod for the small-diameter rods was obtained using MAP with uniform variance and β=0.4. This setting led to RC rod,1 mm=0.21, RC rod,2 mm=0.57, %STD unif=1.38, SOR wat=0.0011, and SOR air=0.00086. However, the highest activity recovery for the smallest rods with still very small %STD unif was obtained using β=0.075, for which these IQ parameters were 0.31, 0.74, 2.67, 0.0041, and 0.0030, respectively. The different settings of reconstruction parameters were clearly reflected in the rat and mouse images as the trade-off between the recovery of

  3. Use of an in-field-of-view shield to improve count rate performance of the single crystal layer high-resolution research tomograph PET scanner for small animal brain scans.

    PubMed

    Boellaard, R; de Jong, H W A M; Molthoff, C F M; Buijs, F; Lenox, M; Nutt, R; Lammertsma, A A

    2003-12-01

    The count rate performance of the single LSO crystal layer high-resolution research tomograph (HRRT-S) PET scanner is limited by the processing speed of its electronics. Therefore, the feasibility of using an in-field-of-view (in-FOV) shield to improve the noise equivalent count rates (NECR) for small animal brain studies was investigated. The in-FOV shield consists of a lead tube of 12 cm length, 6 cm inner diameter and 9 mm wall thickness. It is large enough to shield the activity in the body of a rat or mouse. First, the effect of this shield on NECR was studied. Secondly, a number of experiments were performed to assess the effects of the shield on the accuracy of transmission scan data and, next, on reconstructed activity distribution in the brain. For activities below 150 MBq NECR improved only by 5-10%. For higher activities NECR maxima of 1.2E4 cps at 200 MBq and 2.2E4 cps at 370 MBq were found without and with shield, respectively. Listmode data taken without shield, however, were corrupted for activities above 75 MBq due to data overrun problems (time tag losses) of the electronics. When the shield was used data overrun was avoided up to activities of 150 MBq. For the unshielded part of the phantom, transmission scan data were the same with and without shield. The estimated scatter contribution was approximately 8.5% without and 5.5% with shield. Reconstructed emission data showed a difference up to 5% in the unshielded part of the phantom at 5 mm or more from the edge of the shielding. Of this 5% about 3% results from the difference in the uncorrected scatter contribution. In conclusion, an in-FOV shield can be used successfully in an HRRT PET scanner to improve NECR and accuracy of small animal brain studies. The latter is especially important when high activities are required for tracers with low brain uptake or when multiple animals are scanned simultaneously. PMID:14703172

  4. Quantitative PET imaging with the 3T MR-BrainPET

    NASA Astrophysics Data System (ADS)

    Weirich, C.; Scheins, J.; Lohmann, P.; Tellmann, L.; Byars, L.; Michel, C.; Rota Kops, E.; Brenner, D.; Herzog, H.; Shah, N. J.

    2013-02-01

    The new hybrid imaging technology of MR-PET allows for simultaneous acquisition of versatile MRI contrasts and the quantitative metabolic imaging with PET. In order to achieve the quantification of PET images with minimal residual error the application of several corrections is crucial. In this work we present our results on quantification with the 3T MR BrainPET scanner.

  5. Lutetium oxyorthosilicate (LSO) intrinsic activity correction and minimal detectable target activity study for SPECT imaging with a LSO-based animal PET scanner.

    PubMed

    Yao, Rutao; Ma, Tianyu; Shao, Yiping

    2008-08-21

    This work is part of a feasibility study to develop SPECT imaging capability on a lutetium oxyorthosilicate (LSO) based animal PET system. The SPECT acquisition was enabled by inserting a collimator assembly inside the detector ring and acquiring data in singles mode. The same LSO detectors were used for both PET and SPECT imaging. The intrinsic radioactivity of (176)Lu in the LSO crystals, however, contaminates the SPECT data, and can generate image artifacts and introduce quantification error. The objectives of this study were to evaluate the effectiveness of a LSO background subtraction method, and to estimate the minimal detectable target activity (MDTA) of image object for SPECT imaging. For LSO background correction, the LSO contribution in an image study was estimated based on a pre-measured long LSO background scan and subtracted prior to the image reconstruction. The MDTA was estimated in two ways. The empirical MDTA (eMDTA) was estimated from screening the tomographic images at different activity levels. The calculated MDTA (cMDTA) was estimated from using a formula based on applying a modified Currie equation on an average projection dataset. Two simulated and two experimental phantoms with different object activity distributions and levels were used in this study. The results showed that LSO background adds concentric ring artifacts to the reconstructed image, and the simple subtraction method can effectively remove these artifacts-the effect of the correction was more visible when the object activity level was near or above the eMDTA. For the four phantoms studied, the cMDTA was consistently about five times of the corresponding eMDTA. In summary, we implemented a simple LSO background subtraction method and demonstrated its effectiveness. The projection-based calculation formula yielded MDTA results that closely correlate with that obtained empirically and may have predicative value for imaging applications. PMID:18670052

  6. Lutetium oxyorthosilicate (LSO) intrinsic activity correction and minimal detectable target activity study for SPECT imaging with a LSO-based animal PET scanner

    NASA Astrophysics Data System (ADS)

    Yao, Rutao; Ma, Tianyu; Shao, Yiping

    2008-08-01

    This work is part of a feasibility study to develop SPECT imaging capability on a lutetium oxyorthosilicate (LSO) based animal PET system. The SPECT acquisition was enabled by inserting a collimator assembly inside the detector ring and acquiring data in singles mode. The same LSO detectors were used for both PET and SPECT imaging. The intrinsic radioactivity of 176Lu in the LSO crystals, however, contaminates the SPECT data, and can generate image artifacts and introduce quantification error. The objectives of this study were to evaluate the effectiveness of a LSO background subtraction method, and to estimate the minimal detectable target activity (MDTA) of image object for SPECT imaging. For LSO background correction, the LSO contribution in an image study was estimated based on a pre-measured long LSO background scan and subtracted prior to the image reconstruction. The MDTA was estimated in two ways. The empirical MDTA (eMDTA) was estimated from screening the tomographic images at different activity levels. The calculated MDTA (cMDTA) was estimated from using a formula based on applying a modified Currie equation on an average projection dataset. Two simulated and two experimental phantoms with different object activity distributions and levels were used in this study. The results showed that LSO background adds concentric ring artifacts to the reconstructed image, and the simple subtraction method can effectively remove these artifacts—the effect of the correction was more visible when the object activity level was near or above the eMDTA. For the four phantoms studied, the cMDTA was consistently about five times of the corresponding eMDTA. In summary, we implemented a simple LSO background subtraction method and demonstrated its effectiveness. The projection-based calculation formula yielded MDTA results that closely correlate with that obtained empirically and may have predicative value for imaging applications.

  7. Optical scanner

    NASA Technical Reports Server (NTRS)

    Finkel, Mitchell W. (Inventor)

    1987-01-01

    An optical scanner for imaging lines in an object plane onto a linear array in a focal plane either continuously or discretely is described. The scanner consists of a set of four mutually perpendicularly oriented plane corner mirrors which provide a reflecting path that describes a parallelogram. In addition, there is a plane parallel scanning mirror with a front and back reflecting surface located midway between the first and fourth corner mirrors. It is oriented so that in the mid-scan position it is parallel to the first corner mirror, and therefore perpendicular to the fourth corner mirror. As the scan mirror rotates, rays incident from a plurality of lines in the object plane are selectively directed through the optical system arriving at a common intersection on the back surface of the scanning mirror where the rays are colinearly directed toward a lens and then imaged onto the linear array in the focal plane. A set of compensating mirrors may be introduced just before the imaging lens to compensate for a small and generally negligible path difference delta sub l between the axial and marginal rays.

  8. Bottled liquid explosive scanner by near infrared

    NASA Astrophysics Data System (ADS)

    Itozaki, Hideo

    2016-05-01

    A bottled liquid explosive scanner has been developed using near infrared technology for glass or PET bottles and ultrasound technology for metal cans. It has database of near infrared absorbance spectra and sound velocities of various liquids. Scanned liquids can be identified by using this database. This device has been certified by ECAC and installed at Japanese international airport.

  9. Advances in Clinical PET/MRI Instrumentation.

    PubMed

    Herzog, Hans; Lerche, Christoph

    2016-04-01

    In 2010, the first whole-body PET/MRI scanners installed for clinical use were the sequential Philips PET/MRI with PMT-based, TOF-capable technology and the integrated simultaneous Siemens PET/MRI. Avalanche photodiodes as non-magneto-sensitive readout electronics allowed PET integrated within the MRI. The experiences with these scanners showed that improvements of software aspects, such as attenuation correction, were necessary and that efficient protocols combining optimally PET and MRI must be still developed. In 2014, General Electric issued an integrated PET/MRI with SiPM-based PET detectors, allowing TOF-PET. Looking at the MRI components of current PET/MR imaging systems, primary improvements come from sequences and new coils. PMID:26952724

  10. PET/CT: fundamental principles.

    PubMed

    Seemann, Marcus D

    2004-05-28

    Positron emission tomography (PET) facilitates the evaluation of metabolic and molecular characteristics of a wide variety of cancers, but is limited in its ability to visualize anatomical structures. Computed tomography (CT) facilitates the evaluation of anatomical structures of cancers, but can not visualize their metabolic and molecular aspects. Therefore, the combination of PET and CT provides the ability to accurately register metabolic and molecular aspects of disease with anatomical findings, adding further information to the diagnosis and staging of tumors. The recent generation of high performance PET/CT scanners combines a state of the art full-ring 3D PET scanner and a high-end 16-slice CT scanner. In PET/CT scanners, a CT examination is used for attenuation correction of PET images rather than standard transmission scanning using superset 68 Ge sources. This reduces the examination time, but metallic objects and contrast agents that alter the CT image quality and quantitative measurements of standardized uptake values (SUV) may lead to artifacts in the PET images. Hybrid PET/CT imaging will be very important in oncological applications in the decades to come, and possibly for use in cancer screening and cardiac imaging. PMID:15257877

  11. What is Scanner and NonScanner?

    Atmospheric Science Data Center

    2014-12-08

    ... (ERBE ADM). The scanner is designed for regional to large scale analysis, and due to the smaller footprint, the scanner product is able ... The large footprint (1000 km) is designed only for large scale analysis, thus products provide only all-sky data. Because the nonscanner ...

  12. Compact conscious animal positron emission tomography scanner

    DOEpatents

    Schyler, David J.; O'Connor, Paul; Woody, Craig; Junnarkar, Sachin Shrirang; Radeka, Veljko; Vaska, Paul; Pratte, Jean-Francois; Volkow, Nora

    2006-10-24

    A method of serially transferring annihilation information in a compact positron emission tomography (PET) scanner includes generating a time signal for an event, generating an address signal representing a detecting channel, generating a detector channel signal including the time and address signals, and generating a composite signal including the channel signal and similarly generated signals. The composite signal includes events from detectors in a block and is serially output. An apparatus that serially transfers annihilation information from a block includes time signal generators for detectors in a block and an address and channel signal generator. The PET scanner includes a ring tomograph that mounts onto a portion of an animal, which includes opposing block pairs. Each of the blocks in a block pair includes a scintillator layer, detection array, front-end array, and a serial encoder. The serial encoder includes time signal generators and an address signal and channel signal generator.

  13. Recent developments in PET detector technology

    PubMed Central

    Lewellen, Tom K

    2010-01-01

    Positron emission tomography (PET) is a tool for metabolic imaging that has been utilized since the earliest days of nuclear medicine. A key component of such imaging systems is the detector modules—an area of research and development with a long, rich history. Development of detectors for PET has often seen the migration of technologies, originally developed for high energy physics experiments, into prototype PET detectors. Of the many areas explored, some detector designs go on to be incorporated into prototype scanner systems and a few of these may go on to be seen in commercial scanners. There has been a steady, often very diverse development of prototype detectors, and the pace has accelerated with the increased use of PET in clinical studies (currently driven by PET/CT scanners) and the rapid proliferation of pre-clinical PET scanners for academic and commercial research applications. Most of these efforts are focused on scintillator-based detectors, although various alternatives continue to be considered. For example, wire chambers have been investigated many times over the years and more recently various solid-state devices have appeared in PET detector designs for very high spatial resolution applications. But even with scintillators, there have been a wide variety of designs and solutions investigated as developers search for solutions that offer very high spatial resolution, fast timing, high sensitivity and are yet cost effective. In this review, we will explore some of the recent developments in the quest for better PET detector technology. PMID:18695301

  14. Scanner matching optimization

    NASA Astrophysics Data System (ADS)

    Kupers, Michiel; Klingbeil, Patrick; Tschischgale, Joerg; Buhl, Stefan; Hempel, Fritjof

    2009-03-01

    Cost of ownership of scanners for the manufacturing of front end layers is becoming increasingly expensive. The ability to quickly switch the production of a layer to another scanner in case it is down is important. This paper presents a method to match the scanner grids in the most optimal manner so that use of front end scanners in effect becomes interchangeable. A breakdown of the various components of overlay is given and we discuss methods to optimize the matching strategy in the fab. A concern here is how to separate the scanner and process induced effects. We look at the relative contributions of intrafield and interfield errors caused by the scanner and the process. Experimental results of a method to control the scanner grid are presented and discussed. We compare the overlay results before and after optimizing the scanner grids and show that the matching penalty is reduced by 20%. We conclude with some thoughts on the need to correct the remaining matching errors.

  15. Simulation of triple coincidences in PET.

    PubMed

    Cal-González, J; Lage, E; Herranz, E; Vicente, E; Udias, J M; Moore, S C; Park, M-A; Dave, S R; Parot, V; Herraiz, J L

    2015-01-01

    Although current PET scanners are designed and optimized to detect double coincidence events, there is a significant amount of triple coincidences in any PET acquisition. Triple coincidences may arise from causes such as: inter-detector scatter (IDS), random triple interactions (RT), or the detection of prompt gamma rays in coincidence with annihilation photons when non-pure positron-emitting radionuclides are used (β(+)γ events). Depending on the data acquisition settings of the PET scanner, these triple events are discarded or processed as a set of double coincidences if the energy of the three detected events is within the scanner's energy window. This latter option introduces noise in the data, as at most, only one of the possible lines-of-response defined by triple interactions corresponds to the line along which the decay occurred. Several novel works have pointed out the possibility of using triple events to increase the sensitivity of PET scanners or to expand PET imaging capabilities by allowing differentiation between radiotracers labeled with non-pure and pure positron-emitting radionuclides. In this work, we extended the Monte Carlo simulator PeneloPET to assess the proportion of triple coincidences in PET acquisitions and to evaluate their possible applications. We validated the results of the simulator against experimental data acquired with a modified version of a commercial preclinical PET/CT scanner, which was enabled to acquire and process triple-coincidence events. We used as figures of merit the energy spectra for double and triple coincidences and the triples-to-doubles ratio for different energy windows and radionuclides. After validation, the simulator was used to predict the relative quantity of triple-coincidence events in two clinical scanners assuming different acquisition settings. Good agreement between simulations and preclinical experiments was found, with differences below 10% for most of the observables considered. For clinical

  16. Tunable Resonant Scanners

    NASA Astrophysics Data System (ADS)

    Montagu, Jean I.

    1987-01-01

    The most attractive features of resonant scanners are high reliability and eternal life as well as extremely low wobble and jitter. Power consumption is also low, electronic drive is simple, and the device is capable of handling large beams. All of these features are delivered at a low cost in a small package. The resonant scanner's use in numerous high precision applications, however, has been limited because of the difficulty in controlling its phase and resonant frequency. This paper introduces the concept of tunable/controllable resonant scanners, discusses their features, and offers a number of tuning techniques. It describes two angular scanner designs and presents data on tunable range and life tests. It also reviews applications for these new tunable resonant scanners that preserve the desirable features of earlier models while removing the old problems with synchronization or time base flexibility. The three major types of raster scanning applications where the tunable resonant scanner may be of benefit are: 1. In systems with multiple time bases such as multiple scanner networks or with scanners keyed to a common clock (the line frequency or data source) or a machine with multiple resonant scanners. A typical application is image and text transmission, also a printer with a large data base where a buffer is uneconomical. 2. In systems sharing data processing or laser equipment for reasons of cost or capacity, typically multiple work station manufacturing processes or graphic processes. 3. In systems with extremely precise time bases where the frequency stability of conventional scanners cannot be relied upon.

  17. [The PET, Past and Future].

    PubMed

    Fujii, Hirofumi

    2015-01-01

    Positron emission tomography (PET) is a unique nuclear medicine test using positron emitters such as 18F and 11C. In PET tests, various kinds of functional aspects of human bodies can be evaluated by using compounds labeled by these positron emitters. Recently, combined scanners of PET and anatomical imaging modalities such as CT and MRI have been developed and functional information with anatomical location can be easily obtained, increasing the usefulness of PET tests. PET tests are now essential imaging tools to diagnose various kinds of disease with functional abnormalities. In the field of oncology, 18F-fluorodeoxy glucose PET tests are routinely used in clinical practice under health insurance. In the field of neurology, PET tests are actively used to investigate cerebral function by labeled neurotransmitters and so on. Currently, brain PET tests to detect beta-amyloid are applied to the diagnosis of dementia. In the field of cardiology, cardiac perfusion and myocardial metabolism are quantitatively measured by using PET and obtained results have successfully revealed the pathogenesis of intractable cardiac diseases. Future technical advances will enhance the usefulness of PET tests more and more. PMID:26753390

  18. The ADNI PET Core

    PubMed Central

    Jagust, William J.; Bandy, Dan; Chen, Kewei; Foster, Norman L.; Landau, Susan M.; Mathis, Chester A.; Price, Julie C.; Reiman, Eric M.; Skovronsky, Daniel; Koeppe, Robert A.

    2010-01-01

    Background This is a progress report of the Alzheimer's Disease Neuroimaging Initiative (ADNI) PET Core. Methods The Core has supervised the acquisition, quality control, and analysis of longitudinal [18F]fluorodeoxyglucose PET (FDG-PET) data in approximately half of the ADNI cohort. In an “add on” study, approximately 100 subjects also underwent scanning with [11C]PIB-PET for amyloid imaging. The Core developed quality control procedures and standardized image acquisition by developing an imaging protocol that has been widely adopted in academic and pharmaceutical industry studies. Data processing provides users with scans that have identical orientation and resolution characteristics despite acquisition on multiple scanner models. The Core labs have used a number of different approaches to characterize differences between subject groups (AD, MCI, controls), to examine longitudinal change over time in glucose metabolism and amyloid deposition, and to assess the use of FDG-PET as a potential outcome measure in clinical trials. Results ADNI data indicate that FDG-PET increases statistical power over traditional cognitive measures, might aid subject selection, and could substantially reduce the sample size in a clinical trial. PIB-PET data showed expected group differences, and identified subjects with significant annual increases in amyloid load across the subject groups. The next activities of the PET core in ADNI will entail developing standardized protocols for amyloid imaging using the [18F]-labeled amyloid imaging agent AV45, which can be delivered to virtually all ADNI sites. Conclusions ADNI has demonstrated the feasibility and utility of multicenter PET studies and is helping to clarify the role of biomarkers in the study of aging and dementia. PMID:20451870

  19. Investigation of Holographic Scanners

    NASA Astrophysics Data System (ADS)

    Xiang, Lian Qin

    Holographic scanners are capable of challenging both the speed and resolution of polygon scanners. This work investigates, in detail, the design and operation of a holographic scanner with an aspherical reflector. The characteristics of this holographic scanner are presented through theoretical analyses and computer simulation. The calculated data and the experimental results show that this system has excellent scan line straightness and scan linearity. The influence of the eccentricity and wobble of the hologram on the quality of the scan lines can be minimized by proper choice of system parameters. This unique system can readily perform 1-D, 2 -D, 3-D and selective scans. These features make suitable applications for robot vision, part inspection, high speed printing, and input/output devices for computers. If the hologram is operating in the reflective mode, there are no transmissive components in this scanner. It can be used with acoustic waves and electromagnetic waves with longer wavelengths, such as infrared, microwaves, millimeter waves. Since it is difficult to find a suitable recording material for these waves, a technique for making computer -generated holograms has also been developed here. The practical considerations for making quality holograms are summarized. An improved coating process for photoresist and a novel anti-reflection setup for the hologram plate are developed. The detailed experimental processes are included. The planar grating scanner for one dimensional, two-dimensional and cross-scanning patterns is analyzed and demonstrated. A comparison is made with two other two-dimensional scanners.

  20. PET/CT in radiation oncology

    SciTech Connect

    Pan, Tinsu; Mawlawi, Osama

    2008-11-15

    PET/CT is an effective tool for the diagnosis, staging and restaging of cancer patients. It combines the complementary information of functional PET images and anatomical CT images in one imaging session. Conventional stand-alone PET has been replaced by PET/CT for improved patient comfort, patient throughput, and most importantly the proven clinical outcome of PET/CT over that of PET and that of separate PET and CT. There are over two thousand PET/CT scanners installed worldwide since 2001. Oncology is the main application for PET/CT. Fluorine-18 deoxyglucose is the choice of radiopharmaceutical in PET for imaging the glucose uptake in tissues, correlated with an increased rate of glycolysis in many tumor cells. New molecular targeted agents are being developed to improve the accuracy of targeting different disease states and assessing therapeutic response. Over 50% of cancer patients receive radiation therapy (RT) in the course of their disease treatment. Clinical data have demonstrated that the information provided by PET/CT often changes patient management of the patient and/or modifies the RT plan from conventional CT simulation. The application of PET/CT in RT is growing and will become increasingly important. Continuing improvement of PET/CT instrumentation will also make it easier for radiation oncologists to integrate PET/CT in RT. The purpose of this article is to provide a review of the current PET/CT technology, to project the future development of PET and CT for PET/CT, and to discuss some issues in adopting PET/CT in RT and potential improvements in PET/CT simulation of the thorax in radiation therapy.

  1. Use of PET/CT scanning in cancer patients: technical and practical considerations

    PubMed Central

    2005-01-01

    This overview of the oncologic applications of positron emission tomography (PET) focuses on the technical aspects and clinical applications of a newer technique: the combination of a PET scanner and a computed tomography (CT) scanner in a single (PET/CT) device. Examples illustrate how PET/CT contributes to patient care and improves upon the previous state-of-the-art method of comparing a PET scan with a separate CT scan. Finally, the author presents some of the results from studies of PET/CT imaging that are beginning to appear in the literature. PMID:16252023

  2. Forensics for flatbed scanners

    NASA Astrophysics Data System (ADS)

    Gloe, Thomas; Franz, Elke; Winkler, Antje

    2007-02-01

    Within this article, we investigate possibilities for identifying the origin of images acquired with flatbed scanners. A current method for the identification of digital cameras takes advantage of image sensor noise, strictly speaking, the spatial noise. Since flatbed scanners and digital cameras use similar technologies, the utilization of image sensor noise for identifying the origin of scanned images seems to be possible. As characterization of flatbed scanner noise, we considered array reference patterns and sensor line reference patterns. However, there are particularities of flatbed scanners which we expect to influence the identification. This was confirmed by extensive tests: Identification was possible to a certain degree, but less reliable than digital camera identification. In additional tests, we simulated the influence of flatfielding and down scaling as examples for such particularities of flatbed scanners on digital camera identification. One can conclude from the results achieved so far that identifying flatbed scanners is possible. However, since the analyzed methods are not able to determine the image origin in all cases, further investigations are necessary.

  3. Factors affecting accuracy and precision in PET volume imaging

    SciTech Connect

    Karp, J.S.; Daube-Witherspoon, M.E.; Muehllehner, G. )

    1991-03-01

    Volume imaging positron emission tomographic (PET) scanners with no septa and a large axial acceptance angle offer several advantages over multiring PET scanners. A volume imaging scanner combines high sensitivity with fine axial sampling and spatial resolution. The fine axial sampling minimizes the partial volume effect, which affects the measured concentration of an object. Even if the size of an object is large compared to the slice spacing in a multiring scanner, significant variation in the concentration is measured as a function of the axial position of the object. With a volume imaging scanner, it is necessary to use a three-dimensional reconstruction algorithm in order to avoid variations in the axial resolution as a function of the distance from the center of the scanner. In addition, good energy resolution is needed in order to use a high energy threshold to reduce the coincident scattered radiation.

  4. A proposal of an open PET geometry.

    PubMed

    Yamaya, Taiga; Inaniwa, Taku; Minohara, Shinichi; Yoshida, Eiji; Inadama, Naoko; Nishikido, Fumihiko; Shibuya, Kengo; Lam, Chih Fung; Murayama, Hideo

    2008-02-01

    The long patient port of a PET scanner tends to put stress on patients, especially patients with claustrophobia. It also prevents doctors and technicians from taking care of patients during scanning. In this paper, we proposed an 'open PET' geometry, which consists of two axially separated detector rings. A long and continuous field-of-view (FOV) including a 360 degrees opened gap between two detector rings can be imaged enabling a fully 3D image reconstruction of all the possible lines-of-response. The open PET will become practical if iterative image reconstruction methods are applied even though image reconstruction of the open PET is analytically an incomplete problem. First we implemented a 'masked' 3D ordered subset expectation maximization (OS-EM) in which the system matrix was obtained from a long 'gapless' scanner by applying a mask to detectors corresponding to the open space. Next, in order to evaluate imaging performance of the proposed open PET geometry, we simulated a dual HR+ scanner (ring diameter of D = 827 mm, axial length of W = 154 mm x 2) separated by a variable gap. The gap W was the maximum limit to have axially continuous FOV of 3W though the maximum diameter of FOV at the central slice was limited to D/2. Artifacts, observed on both sides of the open space when the gap exceeded W, were effectively reduced by inserting detectors partially into unnecessary open spaces. We also tested the open PET geometry using experimental data obtained by the jPET-D4. The jPET-D4 is a prototype brain scanner, which has 5 rings of 24 detector blocks. We simulated the open jPET-D4 with a gap of 66 mm by eliminating 1 block-ring from experimental data. Although some artifacts were seen at both ends of the opened gap, very similar images were obtained with and without the gap. The proposed open PET geometry is expected to lead to realization of in-beam PET, which is a method for an in situ monitoring of charged particle therapy, by letting the beams pass

  5. Portable biochip scanner device

    DOEpatents

    Perov, Alexander; Sharonov, Alexei; Mirzabekov, Andrei D.

    2002-01-01

    A portable biochip scanner device used to detect and acquire fluorescence signal data from biological microchips (biochips) is provided. The portable biochip scanner device employs a laser for emitting an excitation beam. An optical fiber delivers the laser beam to a portable biochip scanner. A lens collimates the laser beam, the collimated laser beam is deflected by a dichroic mirror and focused by an objective lens onto a biochip. The fluorescence light from the biochip is collected and collimated by the objective lens. The fluorescence light is delivered to a photomultiplier tube (PMT) via an emission filter and a focusing lens. The focusing lens focuses the fluorescence light into a pinhole. A signal output of the PMT is processed and displayed.

  6. Biochip scanner device

    DOEpatents

    Perov, Alexander; Belgovskiy, Alexander I.; Mirzabekov, Andrei D.

    2001-01-01

    A biochip scanner device used to detect and acquire fluorescence signal data from biological microchips or biochips and method of use are provided. The biochip scanner device includes a laser for emitting a laser beam. A modulator, such as an optical chopper modulates the laser beam. A scanning head receives the modulated laser beam and a scanning mechanics coupled to the scanning head moves the scanning head relative to the biochip. An optical fiber delivers the modulated laser beam to the scanning head. The scanning head collects the fluorescence light from the biochip, launches it into the same optical fiber, which delivers the fluorescence into a photodetector, such as a photodiode. The biochip scanner device is used in a row scanning method to scan selected rows of the biochip with the laser beam size matching the size of the immobilization site.

  7. PET/MRI: challenges, solutions and perspectives.

    PubMed

    Herzog, Hans

    2012-12-01

    Already from the start of PET/CT integrating positron emission tomography (PET) and computed tomography (CT) in one instrument, there have been considerations how to combine PET and magnetic resonance imaging (MRI) so that their complementary abilities can be utilized in a single investigation. Since classical PET electronics fail in an even weak magnetic field and PET signal processing might disturb high-frequency signals of MRI, it soon became clear that new solutions had to be found to avoid mutual interferences. During the last fifteen years a number of different approaches towards PET/MRI for small animal imaging have been developed by research groups which together with their specific features are summarized in this review. Recently, PET/MRI for human imaging became available as well - this time by industrial initiatives. First some prototypes of BrainPET/MRI were developed followed by commercial products for simultaneous and non-simultaneous whole-body PET/MRI. Although only PET/MRI integrated in one scanner offers the full diversity of complementary multiparametric imaging, there are also promising applications of non-simultaneous sequential PET/MRI. While describing the present instrumentation for human PET/MRI, this review discusses the challenges and promises related to this new imaging technology. PMID:22925652

  8. Hybrid registration of PET/CT in thoracic region with pre-filtering PET sinogram

    NASA Astrophysics Data System (ADS)

    Mokri, S. S.; Saripan, M. I.; Marhaban, M. H.; Nordin, A. J.; Hashim, S.

    2015-11-01

    The integration of physiological (PET) and anatomical (CT) images in cancer delineation requires an accurate spatial registration technique. Although hybrid PET/CT scanner is used to co-register these images, significant misregistrations exist due to patient and respiratory/cardiac motions. This paper proposes a hybrid feature-intensity based registration technique for hybrid PET/CT scanner. First, simulated PET sinogram was filtered with a 3D hybrid mean-median before reconstructing the image. The features were then derived from the segmented structures (lung, heart and tumor) from both images. The registration was performed based on modified multi-modality demon registration with multiresolution scheme. Apart from visual observations improvements, the proposed registration technique increased the normalized mutual information index (NMI) between the PET/CT images after registration. All nine tested datasets show marked improvements in mutual information (MI) index than free form deformation (FFD) registration technique with the highest MI increase is 25%.

  9. Professor Pet.

    ERIC Educational Resources Information Center

    Pet Information Bureau, New York, NY.

    This manual outlines ways in which observation and care of classroom pet animals may be used to enrich the education of elementary school children. Part one deals with the benefits of having pets in the classroom. Part two illustrates ways in which pets can serve as valuable teaching tools and gives examples of lessons in which the use of pets can…

  10. Hybrid Dispersion Laser Scanner

    PubMed Central

    Goda, K.; Mahjoubfar, A.; Wang, C.; Fard, A.; Adam, J.; Gossett, D. R.; Ayazi, A.; Sollier, E.; Malik, O.; Chen, E.; Liu, Y.; Brown, R.; Sarkhosh, N.; Di Carlo, D.; Jalali, B.

    2012-01-01

    Laser scanning technology is one of the most integral parts of today's scientific research, manufacturing, defense, and biomedicine. In many applications, high-speed scanning capability is essential for scanning a large area in a short time and multi-dimensional sensing of moving objects and dynamical processes with fine temporal resolution. Unfortunately, conventional laser scanners are often too slow, resulting in limited precision and utility. Here we present a new type of laser scanner that offers ∼1,000 times higher scan rates than conventional state-of-the-art scanners. This method employs spatial dispersion of temporally stretched broadband optical pulses onto the target, enabling inertia-free laser scans at unprecedented scan rates of nearly 100 MHz at 800 nm. To show our scanner's broad utility, we use it to demonstrate unique and previously difficult-to-achieve capabilities in imaging, surface vibrometry, and flow cytometry at a record 2D raster scan rate of more than 100 kHz with 27,000 resolvable points. PMID:22685627

  11. Optical fuel pin scanner

    DOEpatents

    Kirchner, Tommy L.; Powers, Hurshal G.

    1983-01-01

    An optical scanner for indicia arranged in a focal plane at a cylindrical outside surface by use of an optical system including a rotatable dove prism. The dove prism transmits a rotating image of an encircled cylindrical surface area to a stationary photodiode array.

  12. Freestanding Complex Optical Scanners.

    ERIC Educational Resources Information Center

    Frisbie, David A.

    A complex freestanding optical mark recognition (OMR) scanner is one which is not on-line to an external processor; it has intelligence stemming from an internal processor located within the unit or system. The advantages and disadvantages of a complex OMR can best be assessed after identifying the scanning needs and constraints of the potential…

  13. ImmunoPET In Cancer Models

    PubMed Central

    Reddy, Smitha; Robinson, Matthew

    2010-01-01

    Positron Emission Tomography (PET) is playing an increasingly important role in the diagnosis, staging, and monitoring response to treatment in a variety of cancers. Recent efforts have focused on ImmunoPET, which employs antibody-based radiotracers, to image tumors based on expression of tumor-associated antigens. It is postulated that the specificity afforded by antibody targeting should both improve tumor detection and provide phenotypic information related to primary and metastatic lesions that will guide therapy decisions. Advances in antibody-engineering are providing the tools to develop antibody-based molecules with pharmacokinetic properties optimized for use as immunoPET radiotracers. Coupled with technical advances in the design of PET scanners, immunoPET holds promise to improve diagnostic imaging and to guide the use of targeted therapies. An overview of the preclinical immunoPET studies in cancer models is reviewed here. PMID:20350627

  14. What Scanner products are available?

    Atmospheric Science Data Center

    2014-12-08

    ... There are single satellite and combined-satellite scanner products. The best source for these data is to order the ERBE scanner CD which gives all the S4G monthly mean 2.5 degree gridded data from ...

  15. TOPICAL REVIEW: Recent developments in PET detector technology

    NASA Astrophysics Data System (ADS)

    Lewellen, Tom K.

    2008-09-01

    Positron emission tomography (PET) is a tool for metabolic imaging that has been utilized since the earliest days of nuclear medicine. A key component of such imaging systems is the detector modules—an area of research and development with a long, rich history. Development of detectors for PET has often seen the migration of technologies, originally developed for high energy physics experiments, into prototype PET detectors. Of the many areas explored, some detector designs go on to be incorporated into prototype scanner systems and a few of these may go on to be seen in commercial scanners. There has been a steady, often very diverse development of prototype detectors, and the pace has accelerated with the increased use of PET in clinical studies (currently driven by PET/CT scanners) and the rapid proliferation of pre-clinical PET scanners for academic and commercial research applications. Most of these efforts are focused on scintillator-based detectors, although various alternatives continue to be considered. For example, wire chambers have been investigated many times over the years and more recently various solid-state devices have appeared in PET detector designs for very high spatial resolution applications. But even with scintillators, there have been a wide variety of designs and solutions investigated as developers search for solutions that offer very high spatial resolution, fast timing, high sensitivity and are yet cost effective. In this review, we will explore some of the recent developments in the quest for better PET detector technology.

  16. Advances in multimodality imaging through a hybrid PET/MRI system.

    PubMed

    Fatemi-Ardekani, Ali; Samavati, Navid; Tang, Jin; Kamath, Markad V

    2009-01-01

    The development of integrated imaging systems for magnetic resonance imaging (MRI) and positron emission tomography (PET) is currently being explored in a number of laboratories and industrial settings. PET/MRI scanners for both preclinical and human research applications are being developed. PET/MRI overcomes many limitations of PET/computed tomography (CT), such as limited tissue contrast and high radiation doses delivered to the patient or the animal being studied. In addition, recent PET/MRI designs allow for simultaneous rather than sequential acquisition of PET and MRI data, which could not have been achieved through a combination of PET and CT scanners. In a combined PET/CT scanner, while both scanners share a common patient bed, they are hard-wired back-to-back and therefore do not allow simultaneous data acquisition. While PET/MRI offers the possibility of novel imaging strategies, it also creates considerable challenges for acquiring artifact-free images from both modalities. In this review, we discuss motivations, challenges, and potential research applications of developing PET/MRI technology. A brief overview of both MRI and PET is presented and preclinical and clinical applications of PET/MRI are identified. Finally, issues and concerns about image quality, clinical practice, and economic feasibility are discussed. PMID:20565381

  17. Advances in SPECT and PET Hardware.

    PubMed

    Slomka, Piotr J; Pan, Tinsu; Berman, Daniel S; Germano, Guido

    2015-01-01

    There have been significant recent advances in single photon emission computed tomography (SPECT) and positron emission tomography (PET) hardware. Novel collimator designs, such as multi-pinhole and locally focusing collimators arranged in geometries that are optimized for cardiac imaging have been implemented to reduce imaging time and radiation dose. These new collimators have been coupled with solid state photon detectors to further improve image quality and reduce scanner size. The new SPECT scanners demonstrate up to a 7-fold increase in photon sensitivity and up to 2 times improvement in image resolution. Although PET scanners are used primarily for oncological imaging, cardiac imaging can benefit from the improved PET sensitivity of 3D systems without inter-plane septa and implementation of the time-of-flight reconstruction. Additionally, resolution recovery techniques are now implemented by all major PET vendors. These new methods improve image contrast, image resolution, and reduce image noise. Simultaneous PET/magnetic resonance (MR) hybrid systems have been developed. Solid state detectors with avalanche photodiodes or digital silicon photomultipliers have also been utilized in PET. These new detectors allow improved image resolution, higher count rate, as well as a reduced sensitivity to electromagnetic MR fields. PMID:25721706

  18. PET/MRI and PET/CT in Lung Lesions and Thoracic Malignancies.

    PubMed

    Flechsig, Paul; Mehndiratta, Amit; Haberkorn, Uwe; Kratochwil, Clemens; Giesel, Frederik L

    2015-07-01

    More than one decade ago, introduction of integrated PET/CT scanners changed oncologic imaging and oncologic patient management profoundly. With these systems, the metabolic information acquired by PET can be anatomically localized even to small structures such as small primary tumors, lymph nodes, and soft tissue masses owing to the high-resolution multidetector CT scanners. This has made PET/CT a most reliable method for tumor detection, characterization, staging, and response monitoring. The importance of an integrated functional and morphologic approach to better understand the biology of oncologic disease and to improve therapy planning is underlined by the increasing number of PET/CT systems worldwide, leading to an increasing number of scientific publications in the field. The paradigmatic indication of integrated PET/CT is staging of patients with lung cancer, as PET/CT allows for precise pretherapeutic staging and also posttreatment restaging according to the TNM criteria. The growing numbers of targeted therapy strategies in the fields of surgery, chemotherapy, and radiation therapy, which are adapted to dedicated tumor stages, require the exact classifications of each patient's tumor stage. In this context, whole-body examinations using integrated (18)F-FDG-PET/CT have been shown to reduce the side effects of futile invasive procedures and reduce additional costly staging procedures. In this review article, the diagnostic and therapeutic effects of PET/CT examinations are highlighted and compared with some competitive techniques such as scintigraphy, MRI, and, where possible, integrated PET/MRI. PMID:26050655

  19. Competitive Advantage of PET/MRI

    PubMed Central

    Jadvar, Hossein; Colletti, Patrick M.

    2013-01-01

    Multimodality imaging has made great strides in the imaging evaluation of patients with a variety of diseases. Positron emission tomography/computed tomography (PET/CT) is now established as the imaging modality of choice in many clinical conditions, particularly in oncology. While the initial development of combined PET/magnetic resonance imaging (PET/MRI) was in the preclinical arena, hybrid PET/MR scanners are now available for clinical use. PET/MRI combines the unique features of MRI including excellent soft tissue contrast, diffusion-weighted imaging, dynamic contrast-enhanced imaging, fMRI and other specialized sequences as well as MR spectroscopy with the quantitative physiologic information that is provided by PET. Most evidence for the potential clinical utility of PET/MRI is based on studies performed with side-by-side comparison or software-fused MRI and PET images. Data on distinctive utility of hybrid PET/MRI are rapidly emerging. There are potential competitive advantages of PET/MRI over PET/CT. In general, PET/MRI may be preferred over PET/CT where the unique features of MRI provide more robust imaging evaluation in certain clinical settings. The exact role and potential utility of simultaneous data acquisition in specific research and clinical settings will need to be defined. It may be that simultaneous PET/MRI will be best suited for clinical situations that are disease-specific, organ-specific, related to diseases of the children or in those patients undergoing repeated imaging for whom cumulative radiation dose must be kept as low as reasonably achievable. PET/MRI also offers interesting opportunities for use of dual modality probes. Upon clear definition of clinical utility, other important and practical issues related to business operational model, clinical workflow and reimbursement will also be resolved. PMID:23791129

  20. Competitive advantage of PET/MRI.

    PubMed

    Jadvar, Hossein; Colletti, Patrick M

    2014-01-01

    Multimodality imaging has made great strides in the imaging evaluation of patients with a variety of diseases. Positron emission tomography/computed tomography (PET/CT) is now established as the imaging modality of choice in many clinical conditions, particularly in oncology. While the initial development of combined PET/magnetic resonance imaging (PET/MRI) was in the preclinical arena, hybrid PET/MR scanners are now available for clinical use. PET/MRI combines the unique features of MRI including excellent soft tissue contrast, diffusion-weighted imaging, dynamic contrast-enhanced imaging, fMRI and other specialized sequences as well as MR spectroscopy with the quantitative physiologic information that is provided by PET. Most evidence for the potential clinical utility of PET/MRI is based on studies performed with side-by-side comparison or software-fused MRI and PET images. Data on distinctive utility of hybrid PET/MRI are rapidly emerging. There are potential competitive advantages of PET/MRI over PET/CT. In general, PET/MRI may be preferred over PET/CT where the unique features of MRI provide more robust imaging evaluation in certain clinical settings. The exact role and potential utility of simultaneous data acquisition in specific research and clinical settings will need to be defined. It may be that simultaneous PET/MRI will be best suited for clinical situations that are disease-specific, organ-specific, related to diseases of the children or in those patients undergoing repeated imaging for whom cumulative radiation dose must be kept as low as reasonably achievable. PET/MRI also offers interesting opportunities for use of dual modality probes. Upon clear definition of clinical utility, other important and practical issues related to business operational model, clinical workflow and reimbursement will also be resolved. PMID:23791129

  1. LIGA Scanner Control Software

    Energy Science and Technology Software Center (ESTSC)

    1999-02-01

    The LIGA Scanner Software is a graphical user interface package that facilitates controlling the scanning operation of x-rays from a synchrotron and sample manipulation for making LIGA parts. The process requires scanning of the LIGA mask and the PMMA resist through a stationary x-ray beam to provide an evenly distributed x-ray exposure over the wafer. This software package has been written specifically to interface with Aerotech motor controllers.

  2. High throughput optical scanner

    DOEpatents

    Basiji, David A.; van den Engh, Gerrit J.

    2001-01-01

    A scanning apparatus is provided to obtain automated, rapid and sensitive scanning of substrate fluorescence, optical density or phosphorescence. The scanner uses a constant path length optical train, which enables the combination of a moving beam for high speed scanning with phase-sensitive detection for noise reduction, comprising a light source, a scanning mirror to receive light from the light source and sweep it across a steering mirror, a steering mirror to receive light from the scanning mirror and reflect it to the substrate, whereby it is swept across the substrate along a scan arc, and a photodetector to receive emitted or scattered light from the substrate, wherein the optical path length from the light source to the photodetector is substantially constant throughout the sweep across the substrate. The optical train can further include a waveguide or mirror to collect emitted or scattered light from the substrate and direct it to the photodetector. For phase-sensitive detection the light source is intensity modulated and the detector is connected to phase-sensitive detection electronics. A scanner using a substrate translator is also provided. For two dimensional imaging the substrate is translated in one dimension while the scanning mirror scans the beam in a second dimension. For a high throughput scanner, stacks of substrates are loaded onto a conveyor belt from a tray feeder.

  3. Pet Health

    MedlinePlus

    ... Before getting a pet, think carefully about which animal is best for your family. What is each ... Does anyone have pet allergies? What type of animal suits your lifestyle and budget? Once you own ...

  4. Small-animal PET: what is it, and why do we need it?

    PubMed

    Yao, Rutao; Lecomte, Roger; Crawford, Elpida S

    2012-09-01

    Small-animal PET refers to imaging of animals such as rats and mice using dedicated PET scanners. Small-animal PET has been used extensively in modern biomedical research. It provides a quantitative measure of the 3-dimensional distribution of a radiopharmaceutical administered to a live subject noninvasively. In this article, we will discuss the operational and technical aspects of small-animal PET; make some comparisons between small-animal PET and human PET systems; identify the challenges of, opportunities for, and ultimate limitations in applying small-animal PET; and discuss some representative small-animal PET applications. Education objectives: After reading this article, the technologist will be able to explain the requirements and benefits of small-animal PET in biomedical research, describe the design and general characteristics of a small-animal PET system, list and describe some of the challenges of imaging small animals, and discuss several small-animal PET applications. PMID:22582006

  5. Joint PET-MR respiratory motion models for clinical PET motion correction.

    PubMed

    Manber, Richard; Thielemans, Kris; Hutton, Brian F; Wan, Simon; McClelland, Jamie; Barnes, Anna; Arridge, Simon; Ourselin, Sébastien; Atkinson, David

    2016-09-01

    Patient motion due to respiration can lead to artefacts and blurring in positron emission tomography (PET) images, in addition to quantification errors. The integration of PET with magnetic resonance (MR) imaging in PET-MR scanners provides complementary clinical information, and allows the use of high spatial resolution and high contrast MR images to monitor and correct motion-corrupted PET data. In this paper we build on previous work to form a methodology for respiratory motion correction of PET data, and show it can improve PET image quality whilst having minimal impact on clinical PET-MR protocols. We introduce a joint PET-MR motion model, using only 1 min per PET bed position of simultaneously acquired PET and MR data to provide a respiratory motion correspondence model that captures inter-cycle and intra-cycle breathing variations. In the model setup, 2D multi-slice MR provides the dynamic imaging component, and PET data, via low spatial resolution framing and principal component analysis, provides the model surrogate. We evaluate different motion models (1D and 2D linear, and 1D and 2D polynomial) by computing model-fit and model-prediction errors on dynamic MR images on a data set of 45 patients. Finally we apply the motion model methodology to 5 clinical PET-MR oncology patient datasets. Qualitative PET reconstruction improvements and artefact reduction are assessed with visual analysis, and quantitative improvements are calculated using standardised uptake value (SUV(peak) and SUV(max)) changes in avid lesions. We demonstrate the capability of a joint PET-MR motion model to predict respiratory motion by showing significantly improved image quality of PET data acquired before the motion model data. The method can be used to incorporate motion into the reconstruction of any length of PET acquisition, with only 1 min of extra scan time, and with no external hardware required. PMID:27524409

  6. 51. View of upper radar scanner switch in radar scanner ...

    Library of Congress Historic Buildings Survey, Historic Engineering Record, Historic Landscapes Survey

    51. View of upper radar scanner switch in radar scanner building 105 from upper catwalk level showing emanating waveguides from upper switch (upper one-fourth of photograph) and emanating waveguides from lower radar scanner switch in vertical runs. - Clear Air Force Station, Ballistic Missile Early Warning System Site II, One mile west of mile marker 293.5 on Parks Highway, 5 miles southwest of Anderson, Anderson, Denali Borough, AK

  7. Integrated display scanner

    DOEpatents

    Veligdan, James T.

    2004-12-21

    A display scanner includes an optical panel having a plurality of stacked optical waveguides. The waveguides define an inlet face at one end and a screen at an opposite end, with each waveguide having a core laminated between cladding. A projector projects a scan beam of light into the panel inlet face for transmission from the screen as a scan line to scan a barcode. A light sensor at the inlet face detects a return beam reflected from the barcode into the screen. A decoder decodes the return beam detected by the sensor for reading the barcode. In an exemplary embodiment, the optical panel also displays a visual image thereon.

  8. Development of PET/MRI with insertable PET for simultaneous PET and MR imaging of human brain

    SciTech Connect

    Jung, Jin Ho; Choi, Yong Jung, Jiwoong; Kim, Sangsu; Lim, Hyun Keong; Im, Ki Chun; Oh, Chang Hyun; Park, Hyun-wook; Kim, Kyung Min; Kim, Jong Guk

    2015-05-15

    Purpose: The purpose of this study was to develop a dual-modality positron emission tomography (PET)/magnetic resonance imaging (MRI) with insertable PET for simultaneous PET and MR imaging of the human brain. Methods: The PET detector block was composed of a 4 × 4 matrix of detector modules, each consisting of a 4 × 4 array LYSO coupled to a 4 × 4 Geiger-mode avalanche photodiode (GAPD) array. The PET insert consisted of 18 detector blocks, circularly mounted on a custom-made plastic base to form a ring with an inner diameter of 390 mm and axial length of 60 mm. The PET gantry was shielded with gold-plated conductive fabric tapes with a thickness of 0.1 mm. The charge signals of PET detector transferred via 4 m long flat cables were fed into the position decoder circuit. The flat cables were shielded with a mesh-type aluminum sheet with a thickness of 0.24 mm. The position decoder circuit and field programmable gate array-embedded DAQ modules were enclosed in an aluminum box with a thickness of 10 mm and located at the rear of the MR bore inside the MRI room. A 3-T human MRI system with a Larmor frequency of 123.7 MHz and inner bore diameter of 60 cm was used as the PET/MRI hybrid system. A custom-made radio frequency (RF) coil with an inner diameter of 25 cm was fabricated. The PET was positioned between gradient and the RF coils. PET performance was measured outside and inside the MRI scanner using echo planar imaging, spin echo, turbo spin echo, and gradient echo sequences. MRI performance was also evaluated with and without the PET insert. The stability of the newly developed PET insert was evaluated and simultaneous PET and MR images of a brain phantom were acquired. Results: No significant degradation of the PET performance caused by MR was observed when the PET was operated using various MR imaging sequences. The signal-to-noise ratio of MR images was slightly degraded due to the PET insert installed inside the MR bore while the homogeneity was

  9. PET/MRI: THE NEXT GENERATION OF MULTI-MODALITY IMAGING?

    PubMed Central

    Pichler, Bernd; Wehrl, Hans F; Kolb, Armin; Judenhofer, Martin S

    2009-01-01

    Multi-modal imaging is now well-established in routine clinical practice. Especially in the field of Nuclear Medicine, new PET installations are comprised almost exclusively of combined PET/CT scanners rather than PET-only systems. However, PET/CT has certain notable shortcomings, including the inability to perform simultaneous data acquisition and the significant radiation dose to the patient contributed by CT. MRI offers, compared to CT, better contrast among soft tissues as well as functional-imaging capabilities. Therefore, the combination of PET with MRI provides many advantages which go far beyond simply combining functional PET information with structural MRI information. Many technical challenges, including possible interference between these modalities, have to be solved when combining PET and MRI and various approaches have been adapted to resolving these issues. Here we present an overview of current working prototypes of combined PET/MRI scanners from different groups. In addition, besides PET/MR images of mice, the first such images of a rat PET/MR, acquired with the first commercial clinical PET/MRI scanner, are presented. The combination of PET and MR is a promising tool in pre-clinical research and will certainly progress to clinical application. PMID:18396179

  10. PeneloPET, a Monte Carlo PET simulation tool based on PENELOPE: features and validation.

    PubMed

    España, S; Herraiz, J L; Vicente, E; Vaquero, J J; Desco, M; Udias, J M

    2009-03-21

    Monte Carlo simulations play an important role in positron emission tomography (PET) imaging, as an essential tool for the research and development of new scanners and for advanced image reconstruction. PeneloPET, a PET-dedicated Monte Carlo tool, is presented and validated in this work. PeneloPET is based on PENELOPE, a Monte Carlo code for the simulation of the transport in matter of electrons, positrons and photons, with energies from a few hundred eV to 1 GeV. PENELOPE is robust, fast and very accurate, but it may be unfriendly to people not acquainted with the FORTRAN programming language. PeneloPET is an easy-to-use application which allows comprehensive simulations of PET systems within PENELOPE. Complex and realistic simulations can be set by modifying a few simple input text files. Different levels of output data are available for analysis, from sinogram and lines-of-response (LORs) histogramming to fully detailed list mode. These data can be further exploited with the preferred programming language, including ROOT. PeneloPET simulates PET systems based on crystal array blocks coupled to photodetectors and allows the user to define radioactive sources, detectors, shielding and other parts of the scanner. The acquisition chain is simulated in high level detail; for instance, the electronic processing can include pile-up rejection mechanisms and time stamping of events, if desired. This paper describes PeneloPET and shows the results of extensive validations and comparisons of simulations against real measurements from commercial acquisition systems. PeneloPET is being extensively employed to improve the image quality of commercial PET systems and for the development of new ones. PMID:19242053

  11. Multispectral scanner optical system

    NASA Technical Reports Server (NTRS)

    Stokes, R. C.; Koch, N. G. (Inventor)

    1980-01-01

    An optical system for use in a multispectral scanner of the type used in video imaging devices is disclosed. Electromagnetic radiation reflected by a rotating scan mirror is focused by a concave primary telescope mirror and collimated by a second concave mirror. The collimated beam is split by a dichroic filter which transmits radiant energy in the infrared spectrum and reflects visible and near infrared energy. The long wavelength beam is filtered and focused on an infrared detector positioned in a cryogenic environment. The short wavelength beam is dispersed by a pair of prisms, then projected on an array of detectors also mounted in a cryogenic environment and oriented at an angle relative to the optical path of the dispersed short wavelength beam.

  12. Optical Scanner for Linear Arrays

    NASA Technical Reports Server (NTRS)

    Finkel, M. W.

    1986-01-01

    Optical scanner instantaneously reads contiguous lines forming scene or target in object plane. Reading active or passive and scans, continuous or discrete. Scans essentially linear with scan angle and symmetric about axial ray. Nominal focal error, resulting from curvature of scan, well within Rayleigh limit. Scanner specifically designed to be fully compatible with general requirements of linear arrays.

  13. Design Features and Mutual Compatibility Studies of the Time-of-Flight PET Capable GE SIGNA PET/MR System.

    PubMed

    Levin, Craig S; Maramraju, Sri Harsha; Khalighi, Mohammad Mehdi; Deller, Timothy W; Delso, Gaspar; Jansen, Floris

    2016-08-01

    A recent entry into the rapidly evolving field of integrated PET/MR scanners is presented in this paper: a whole body hybrid PET/MR system (SIGNA PET/MR, GE Healthcare) capable of simultaneous acquisition of both time-of-flight (TOF) PET and high resolution MR data. The PET ring was integrated into an existing 3T MR system resulting in a (patient) bore opening of 60 cm diameter, with a 25 cm axial FOV. PET performance was evaluated both on the standalone PET ring and on the same detector integrated into the MR system, to assess the level of mutual interference between both subsystems. In both configurations we obtained detector performance data. PET detector performance was not significantly affected by integration into the MR system. The global energy resolution was within 2% (10.3% versus 10.5%), and the system coincidence time resolution showed a maximum change of < 3% (385 ps versus 394 ps) when measured outside MR and during simultaneous PET/MRI acquisitions, respectively. To evaluate PET image quality and resolution, the NEMA IQ phantom was acquired with MR idle and with MR active. Impact of PET on MR IQ was assessed by comparing SNR with PET acquisition on and off. B0 and B1 homogeneities were acquired before and after the integration of the PET ring inside the magnet. In vivo brain and whole body head-to-thighs data were acquired to demonstrate clinical image quality. PMID:26978664

  14. Space-multiplexed optical scanner.

    PubMed

    Riza, Nabeel A; Yaqoob, Zahid

    2004-05-01

    A low-loss two-dimensional optical beam scanner that is capable of delivering large (e.g., > 10 degrees) angular scans along the elevation as well as the azimuthal direction is presented. The proposed scanner is based on a space-switched parallel-serial architecture that employs a coarse-scanner module and a fine-scanner module that produce an ultrahigh scan space-fill factor, e.g., 900 x 900 distinguishable beams in a 10 degrees (elevation) x 10 degrees (azimuth) scan space. The experimentally demonstrated one-dimensional version of the proposed scanner has a supercontinuous scan, 100 distinguishable beam spots in a 2.29 degrees total scan range, and 1.5-dB optical insertion loss. PMID:15130010

  15. PET scan

    MedlinePlus

    You may feel a sharp sting when the needle with the tracer is placed into your vein. A PET scan causes no pain. The table may be ... The amount of radiation used in a PET scan is about the same amount as used in most CT scans. These scans use ...

  16. Reproducibility of 18F-FDG PET uptake measurements in head and neck squamous cell carcinoma on both PET/CT and PET/MR

    PubMed Central

    Fischer, B M; Aznar, M C; Hansen, A E; Vogelius, I R; Löfgren, J; Andersen, F L; Loft, A; Kjaer, A; Højgaard, L; Specht, L

    2015-01-01

    Objective: To investigate reproducibility of fluorine-18 fludeoxyglucose (18F-FDG) uptake on 18F-FDG positron emission tomography (PET)/CT and 18F-FDG PET/MR scans in patients with head and neck squamous cell carcinoma (HNSCC). Methods: 30 patients with HNSCC were included in this prospective study. The patients were scanned twice before radiotherapy treatment with both PET/CT and PET/MR. Patients were scanned on the same scanners, 3 days apart and according to the same protocol. Metabolic tumour activity was measured by the maximum and peak standardized uptake value (SUVmax and SUVpeak, respectively), and total lesion glycolysis from the metabolic tumour volume defined from ≥50% SUVmax. Bland–Altman analysis with limits of agreement, coefficient of variation (CV) from the two modalities were performed in order to test the reproducibility. Furthermore, CVs from SUVmax and SUVpeak were compared. The area under the curve from cumulative SUV–volume histograms were measured and tested for reproducibility of the distribution of 18F-FDG uptake. Results: 24 patients had two pre-treatment PET/CT scans and 21 patients had two pre-treatment PET/MR scans available for further analyses. Mean difference for SUVmax, peak and mean was approximately 4% for PET/CT and 3% for PET/MR, with 95% limits of agreement less than ±20%. CV was small (5–7%) for both modalities. There was no significant difference in CVs between PET/CT and PET/MR (p = 0.31). SUVmax was not more reproducible than SUVpeak (p = 0.09). Conclusion: 18F-FDG uptake in PET/CT and PET/MR is highly reproducible and we found no difference in reproducibility between PET/CT and PET/MR. Advances in knowledge: This is the first report to test reproducibility of PET/CT and PET/MR. PMID:25634069

  17. New cardiac cameras: single-photon emission CT and PET.

    PubMed

    Slomka, Piotr J; Berman, Daniel S; Germano, Guido

    2014-07-01

    Nuclear cardiology instrumentation has evolved significantly in the recent years. Concerns about radiation dose and long acquisition times have propelled developments of dedicated high-efficiency cardiac SPECT scanners. Novel collimator designs, such as multipinhole or locally focusing collimators arranged in geometries that are optimized for cardiac imaging, have been implemented to enhance photon-detection sensitivity. Some of these new SPECT scanners use solid-state photon detectors instead of photomultipliers to improve image quality and to reduce the scanner footprint. These new SPECT devices allow dramatic up to 7-fold reduction in acquisition times or similar reduction in radiation dose. In addition, new hardware for photon attenuation correction allowing ultralow radiation doses has been offered by some vendors. To mitigate photon attenuation artifacts for the new SPECT scanners not equipped with attenuation correction hardware, 2-position (upright-supine or prone-supine) imaging has been proposed. PET hardware developments have been primarily driven by the requirements of oncologic imaging, but cardiac imaging can benefit from improved PET image quality and improved sensitivity of 3D systems. The time-of-flight reconstruction combined with resolution recovery techniques is now implemented by all major PET vendors. These new methods improve image contrast and image resolution and reduce image noise. High-sensitivity 3D PET without interplane septa allows reduced radiation dose for cardiac perfusion imaging. Simultaneous PET/MR hybrid system has been developed. Solid-state PET detectors with avalanche photodiodes or digital silicon photomultipliers have been introduced, and they offer improved imaging characteristics and reduced sensitivity to electromagnetic MR fields. Higher maximum count rate of the new PET detectors allows routine first-pass Rb-82 imaging, with 3D PET acquisition enabling clinical utilization of dynamic imaging with myocardial flow

  18. Future image acquisition trends for PET/MRI.

    PubMed

    Boss, Andreas; Weiger, Markus; Wiesinger, Florian

    2015-05-01

    Hybrid PET/MRI scanners have become commercially available in the past years but are not yet widely distributed. The combination of a state-of-the-art PET with a state-of-the-art MRI scanner provides numerous potential advantages compared with the established PET/CT hybrid systems, namely, increased soft tissue contrast; functional information from MRI such as diffusion, perfusion, and blood oxygenation level-dependent techniques; true multiplanar data acquisition; and reduced radiation exposure. On the contrary, current PET/MRI technology is hampered by several shortcomings compared with PET/CT, the most important issues being how to use MR data for PET attenuation correction and the low sensitivity of MRI for small-scale pulmonary pathologies compared with high-resolution CT. Moreover, the optimal choice for hybrid PET/MRI acquisition protocols needs to be defined providing the highest possible degree of sensitivity and specificity within the constraints of the available measurement time. A multitude of new acquisition strategies of PET and MRI not only offer to overcome current obstacles of hybrid PET/MRI but also provide deeper insights into the pathophysiology of oncological, inflammatory, or degenerative diseases from the combination of molecular and functional imaging techniques. PMID:25841275

  19. Side scanner for supermarkets: a new scanner design standard

    NASA Astrophysics Data System (ADS)

    Cheng, Charles K.; Cheng, J. K.

    1996-09-01

    High speed UPC bar code has become a standard mode of data capture for supermarkets in the US, Europe, and Japan. The influence of the ergonomics community on the design of the scanner is evident. During the past decade the ergonomic issues of cashier in check-outs has led to occupational hand-wrist cumulative trauma disorders, in most cases causing carpal tunnel syndrome, a permanent hand injury. In this paper, the design of a side scanner to resolve the issues is discussed. The complex optical module and the sensor for aforesaid side scanner is described. The ergonomic advantages offer the old counter mounted vertical scanner has been experimentally proved by the industrial funded study at an independent university.

  20. Coastal Zone Color Scanner

    NASA Technical Reports Server (NTRS)

    Johnson, B.

    1988-01-01

    The Coastal Zone Color Scanner (CZCS) spacecraft ocean color instrument is capable of measuring and mapping global ocean surface chlorophyll concentration. It is a scanning radiometer with multiband capability. With new electronics and some mechanical, and optical re-work, it probably can be made flight worthy. Some additional components of a second flight model are also available. An engineering study and further tests are necessary to determine exactly what effort is required to properly prepare the instrument for spaceflight and the nature of interfaces to prospective spacecraft. The CZCS provides operational instrument capability for monitoring of ocean productivity and currents. It could be a simple, low cost alternative to developing new instruments for ocean color imaging. Researchers have determined that with global ocean color data they can: specify quantitatively the role of oceans in the global carbon cycle and other major biogeochemical cycles; determine the magnitude and variability of annual primary production by marine phytoplankton on a global scale; understand the fate of fluvial nutrients and their possible affect on carbon budgets; elucidate the coupling mechanism between upwelling and large scale patterns in ocean basins; answer questions concerning the large scale distribution and timing of spring blooms in the global ocean; acquire a better understanding of the processes associated with mixing along the edge of eddies, coastal currents, western boundary currents, etc., and acquire global data on marine optical properties.

  1. [Combined PET-MRI of the abdomen].

    PubMed

    Vag, Tibor; Eiber, M; Schwaiger, M

    2015-12-01

    The first fully integrated combined positron emission tomography-magnetic resonance imaging (PET-MRI) scanners have been clinically available since 2010. Large prospective studies regarding indications and diagnostic accuracy of this new modality are not yet available; however, preliminary studies have shown a higher diagnostic accuracy and confidence compared to PET-computed tomography (PET-CT) in regions where MRI is known to be superior to CT, such as the liver. The benefit of MRI in accurate lesion characterization and the additional value of diffusion-weighted imaging (DWI) as a complementary functional modality by means of the apparent diffusion coefficient (ADC) is apparent in entities with low tracer uptake (e.g. due to small size) and a decreased or absent accumulation pattern on PET. PMID:26610681

  2. Experimental assessment of resolution improvement of a zoom-in PET.

    PubMed

    Qi, Jinyi; Yang, Yongfeng; Zhou, Jian; Wu, Yibao; Cherry, Simon R

    2011-09-01

    We have proposed a zoom-in positron emission tomography (PET) system that incorporates a high-resolution detector into an existing PET scanner to obtain high-resolution images of a region of interest. Previously we have shown by computer simulations that the high-resolution detector can improve the overall system performance in terms of spatial resolution and lesion detectability. In this study, we assessed the resolution improvement in a real system by incorporating a high-resolution detector into our existing microPET II scanner. The high-resolution detector consists of a 14 × 28 array of 0.5 × 0.5 × 10 mm³ lutetium oxyorthosilicate scintillator elements and is placed near the center of the microPET II scanner. It is coupled to two 64-channel photomultiplier tubes (PMTs) via tapered optical fiber bundles. The PMT signals were read out by the electronics in the microPET II scanner. A 15 µCi Na-22 point source was positioned at various locations above the high-resolution detector. Images were reconstructed using the data measured by the microPET II scanner alone and the microPET II data combined with the high-resolution detector data. Profiles taken through the reconstructed point sources show substantial reduction in full-width-at-half-maximum along the direction parallel to the face of the high-resolution detector. PMID:21828899

  3. Lung PET scan

    MedlinePlus

    ... emission tomography; PET - chest; PET - lung; PET - tumor imaging ... Grainger & Allison's Diagnostic Radiology: A Textbook of Medical Imaging . 6th ed. Philadelphia, PA: Elsevier Churchill Livingstone; 2015: ...

  4. PET and PET/CT imaging of skeletal metastases

    PubMed Central

    2010-01-01

    Abstract Bone scintigraphy augmented with radiographs or cross-sectional imaging, such as computed tomography (CT) or magnetic resonance imaging (MRI), has remained the commonest method to diagnose and follow up skeletal metastases. However, bone scintigraphy is associated with relatively poor spatial resolution, limited diagnostic specificity and reduced sensitivity for bone marrow disease. It also shows limited diagnostic accuracy in assessing response to therapy in a clinically useful time period. With the advent of hybrid positron emission tomography (PET)/CT scanners there has been an increasing interest in using various PET tracers to evaluate skeletal disease including [18F]fluoride (NaF) as a bone-specific tracer and [18F]fluorodeoxyglucose and [18F]choline as tumour-specific tracers. There is also early work exploring the receptor status of skeletal metastases with somatostatin receptor analogues. This review describes the potential utility of these tracers in the assessment of skeletal metastases. PMID:20663736

  5. MSS D Multispectral Scanner System

    NASA Technical Reports Server (NTRS)

    Lauletta, A. M.; Johnson, R. L.; Brinkman, K. L. (Principal Investigator)

    1982-01-01

    The development and acceptance testing of the 4-band Multispectral Scanners to be flown on LANDSAT D and LANDSAT D Earth resources satellites are summarized. Emphasis is placed on the acceptance test phase of the program. Test history and acceptance test algorithms are discussed. Trend data of all the key performance parameters are included and discussed separately for each of the two multispectral scanner instruments. Anomalies encountered and their resolutions are included.

  6. Michigan experimental multispectral scanner system

    NASA Technical Reports Server (NTRS)

    Hasell, P. G., Jr.

    1972-01-01

    A functional description of a multispectral airborne scanner system that provides spectral bands along a single optical line of sight is reported. The airborne scanner consists of an optical telescope for scanning plane perpendicular to the longitudinal axis of the aircraft and radiation detectors for converting radiation to electrical signals. The system makes a linear transformation of input radiation to voltage recorded on analog magnetic tape.

  7. A new electronic read-out for the YAPPET scanner

    NASA Astrophysics Data System (ADS)

    Damiani, C.; Cotta Ramusino, A.; Malaguti, R.; Del Guerra, A.; Di Domenico, G.; Zavattini, G.

    2002-09-01

    A small animal PET-SPECT scanner (YAPPET) prototype was built at the Physics Department of the Ferrara University and is presently being used at the Nuclear Medicine Department for radiopharmaceutical studies on rats. The first YAPPET prototype shows very good performances, but needs some improvements before it can be fully used for intensive radiopharmaceutical research. The main problem of the actual prototype is its heavy electronics, based on NIM and CAMAC standard modules. For this reason a new, compact read-out electronics was developed and tested. The results of a first series of tests made on the first prototype will be presented in the paper.

  8. 3D ultrafast laser scanner

    NASA Astrophysics Data System (ADS)

    Mahjoubfar, A.; Goda, K.; Wang, C.; Fard, A.; Adam, J.; Gossett, D. R.; Ayazi, A.; Sollier, E.; Malik, O.; Chen, E.; Liu, Y.; Brown, R.; Sarkhosh, N.; Di Carlo, D.; Jalali, B.

    2013-03-01

    Laser scanners are essential for scientific research, manufacturing, defense, and medical practice. Unfortunately, often times the speed of conventional laser scanners (e.g., galvanometric mirrors and acousto-optic deflectors) falls short for many applications, resulting in motion blur and failure to capture fast transient information. Here, we present a novel type of laser scanner that offers roughly three orders of magnitude higher scan rates than conventional methods. Our laser scanner, which we refer to as the hybrid dispersion laser scanner, performs inertia-free laser scanning by dispersing a train of broadband pulses both temporally and spatially. More specifically, each broadband pulse is temporally processed by time stretch dispersive Fourier transform and further dispersed into space by one or more diffractive elements such as prisms and gratings. As a proof-of-principle demonstration, we perform 1D line scans at a record high scan rate of 91 MHz and 2D raster scans and 3D volumetric scans at an unprecedented scan rate of 105 kHz. The method holds promise for a broad range of scientific, industrial, and biomedical applications. To show the utility of our method, we demonstrate imaging, nanometer-resolved surface vibrometry, and high-precision flow cytometry with real-time throughput that conventional laser scanners cannot offer due to their low scan rates.

  9. PET Imaging: Basics and New Trends

    NASA Astrophysics Data System (ADS)

    Dahlbom, Magnus

    Positron Emission Tomography or PET is a noninvasive molecular imaging method used both in research to study biology and disease, and clinically as a routine diagnostic imaging tool. In PET imaging, the subject is injected with a tracer labeled with a positron-emitting isotope and is then placed in a scanner to localize the radioactive tracer in the body. The localization of the tracer utilizes the unique decay characteristics of isotopes decaying by positron emission. In the PET scanner, a large number of scintillation detectors use coincidence detection of the annihilation radiation that is emitted as a result of the positron decay. By collecting a large number of these coincidence events, together with tomographic image reconstruction methods, the 3-D distribution of the radioactive tracer in the body can be reconstructed. Depending on the type of tracer used, the distribution will reflect a particular biological process, such as glucose metabolism when fluoro-deoxyglucose is used. PET has evolved from a relatively inefficient single-slice imaging system with relatively poor spatial resolution to an efficient, high-resolution imaging modality which can acquire a whole-body scan in a few minutes. This chapter will describe the basic physics and instrumentation used in PET. The various corrections that are necessary to apply to the acquired data in order to produce quantitative images are also described. Finally, some of the latest trends in instrumentation development are also discussed.

  10. Cardiac PET Perfusion: Prognosis, Risk Stratification, Clinical Management

    PubMed Central

    Dorbala, Sharmila; Di Carli, Marcelo F.

    2014-01-01

    Myocardial perfusion imaging (MPI) with positron emission tomography (PET) has expanded significantly over the past decade. With the wider availability of PET scanners and the routine use of quantitative blood flow imaging, the clinical use of PET MPI is expected to increase further. PET MPI is a powerful tool to identify risk, to quantify risk, and to guide therapy in patients with known or suspected coronary artery disease (CAD). A large body of evidence supports the prognostic value of PET MPI and ejection fraction in intermediate to high risk subjects, in women, in obese individuals and in post coronary artery bypass grafting (CABG) individuals. A normal perfusion study indicates low risk (< 1% annualized rate of cardiac events of cardiac death and non-fatal myocardial infarction), while an abnormal study indicates high risk. With accurate risk stratification, high quality images, and quantitation PET MPI may transform the management of patients with known or suspected CAD. PMID:25234079

  11. EXPLORER: Changing the molecular imaging paradigm with total-body PET/CT (Conference Presentation)

    NASA Astrophysics Data System (ADS)

    Cherry, Simon R.; Badawi, Ramsey D.; Jones, Terry

    2016-04-01

    Positron emission tomography (PET) is the highest sensitivity technique for human whole-body imaging studies. However, current clinical PET scanners do not make full use of the available signal, as they only permit imaging of a 15-25 cm segment of the body at one time. Given the limited sensitive region, whole-body imaging with clinical PET scanners requires relatively long scan times and subjects the patient to higher than necessary radiation doses. The EXPLORER initiative aims to build a 2-meter axial length PET scanner to allow imaging the entire subject at once, capturing nearly the entire available PET signal. EXPLORER will acquire data with ~40-fold greater sensitivity leading to a six-fold increase in reconstructed signal-to-noise ratio for imaging the total body. Alternatively, total-body images with the EXPLORER scanner will be able to be acquired in ~30 seconds or with ~0.15 mSv injected dose, while maintaining current PET image quality. The superior sensitivity will open many new avenues for biomedical research. Specifically for cancer applications, high sensitivity PET will enable detection of smaller lesions. Additionally, greater sensitivity will allow imaging out to 10 half-lives of positron emitting radiotracers. This will enable 1) metabolic ultra-staging with FDG by extending the uptake and clearance time to 3-5 hours to significantly improve contrast and 2) improved kinetic imaging with short-lived radioisotopes such as C-11, crucial for drug development studies. Frequent imaging studies of the same subject to study disease progression or to track response to therapy will be possible with the low dose capabilities of the EXPLORER scanner. The low dose capabilities will also open up new imaging possibilities in pediatrics and adolescents to better study developmental disorders. This talk will review the basis for developing total-body PET, potential applications, and review progress to date in developing EXPLORER, the first total-body PET scanner.

  12. Positron Emission Tomography (PET)

    SciTech Connect

    Welch, M.J.

    1990-01-01

    Positron emission tomography (PET) assesses biochemical processes in the living subject, producing images of function rather than form. Using PET, physicians are able to obtain not the anatomical information provided by other medical imaging techniques, but pictures of physiological activity. In metaphoric terms, traditional imaging methods supply a map of the body's roadways, its, anatomy; PET shows the traffic along those paths, its biochemistry. This document discusses the principles of PET, the radiopharmaceuticals in PET, PET research, clinical applications of PET, the cost of PET, training of individuals for PET, the role of the United States Department of Energy in PET, and the futures of PET. 22 figs.

  13. Positron Emission Tomography (PET)

    DOE R&D Accomplishments Database

    Welch, M. J.

    1990-01-01

    Positron emission tomography (PET) assesses biochemical processes in the living subject, producing images of function rather than form. Using PET, physicians are able to obtain not the anatomical information provided by other medical imaging techniques, but pictures of physiological activity. In metaphoric terms, traditional imaging methods supply a map of the body's roadways, its, anatomy; PET shows the traffic along those paths, its biochemistry. This document discusses the principles of PET, the radiopharmaceuticals in PET, PET research, clinical applications of PET, the cost of PET, training of individuals for PET, the role of the United States Department of Energy in PET, and the futures of PET.

  14. MRI data driven partial volume effects correction in PET imaging using 3D local multi-resolution analysis

    NASA Astrophysics Data System (ADS)

    Le Pogam, Adrien; Lamare, Frederic; Hatt, Mathieu; Fernandez, Philippe; Le Rest, Catherine Cheze; Visvikis, Dimitris

    2013-02-01

    PET partial volume effects (PVE) resulting from the limited resolution of PET scanners is still a quantitative issue that PET/MRI scanners do not solve by themselves. A recently proposed voxel-based locally adaptive 3D multi-resolution PVE correction based on the mutual analysis of wavelet decompositions was applied on 12 clinical 18F-FLT PET/T1 MRI images of glial tumors, and compared to a PET only voxel-wise iterative deconvolution approach. Quantitative and qualitative results demonstrated the interest of exploiting PET/MRI information with higher uptake increases (19±8% vs. 11±7%, p=0.02), as well as more convincing visual restoration of details within tumors with respect to deconvolution of the PET uptake only. Further studies are now required to demonstrate the accuracy of this restoration with histopathological validation of the uptake in tumors.

  15. MR Imaging-Guided Attenuation Correction of PET Data in PET/MR Imaging.

    PubMed

    Izquierdo-Garcia, David; Catana, Ciprian

    2016-04-01

    Attenuation correction (AC) is one of the most important challenges in the recently introduced combined PET/magnetic resonance (MR) scanners. PET/MR AC (MR-AC) approaches aim to develop methods that allow accurate estimation of the linear attenuation coefficients of the tissues and other components located in the PET field of view. MR-AC methods can be divided into 3 categories: segmentation, atlas, and PET based. This review provides a comprehensive list of the state-of-the-art MR-AC approaches and their pros and cons. The main sources of artifacts are presented. Finally, this review discusses the current status of MR-AC approaches for clinical applications. PMID:26952727

  16. Multispectral Scanner for Monitoring Plants

    NASA Technical Reports Server (NTRS)

    Gat, Nahum

    2004-01-01

    A multispectral scanner has been adapted to capture spectral images of living plants under various types of illumination for purposes of monitoring the health of, or monitoring the transfer of genes into, the plants. In a health-monitoring application, the plants are illuminated with full-spectrum visible and near infrared light and the scanner is used to acquire a reflected-light spectral signature known to be indicative of the health of the plants. In a gene-transfer- monitoring application, the plants are illuminated with blue or ultraviolet light and the scanner is used to capture fluorescence images from a green fluorescent protein (GFP) that is expressed as result of the gene transfer. The choice of wavelength of the illumination and the wavelength of the fluorescence to be monitored depends on the specific GFP.

  17. Does the Novel Integrated PET/MRI Offer the Same Diagnostic Performance as PET/CT for Oncological Indications?

    PubMed Central

    Yin, Dayi; Zhang, Jinming; Chen, Yingmao; An, Ningyu; Xu, Baixuan

    2014-01-01

    Background We compared PET/MRI with PET/CT in terms of lesion detection and quantitative measurement to verify the feasibility of the novel integrated imaging modality for oncological applications. Methodology/Principal Findings In total, 285 patients referred to our PET/CT center for oncological indications voluntarily participated in this same-day PET/CT and PET/MRI comparative study. PET/CT images were acquired and reconstructed following routine protocols, and then PET/MRI was performed at a mean time interval of 28±11 min (range 15–45 min). PET/MRI covered the body trunk with a sequence combination of transverse T1WI 3D-volumetric interpolated breath-hold, T2WI turbo spin echo with fat saturation, diffusion-weighted imaging with double b values (50 and 800 s/mm2), and simultaneous PET acquisition over 45 min/5 bed positions. The maximum standardized uptake value (SUVmax) was assessed by manually drawn regions of interest over fluorodeoxyglucose-positive lesions. Among 285 cases, 57 showed no abnormalities, and 368 lesions (278 malignant, 68 benign and 22 undetermined) were detected in 228 patients. When stand-alone modalities were evaluated, PET revealed 31 and 12 lesions missed by CT and MRI, respectively, and CT and MRI revealed 38 and 61 more lesions, respectively, than PET. Compared to CT, MRI detected 40 more lesions and missed 8. In the integrated mode, PET/CT correctly detected 6 lesions misdiagnosed by PET/MRI, but was false-negative in 30 cases that were detected by PET/MRI. The overall diagnosis did not differ between integrated PET/MRI and PET/CT. SUVmax for lesions were slightly higher from PET/MRI than PET/CT but correlated well (ρ = 0.85–0.91). Conclusions/Significance The novel integrated PET/MRI performed comparatively to PET/CT in lesion detection and quantitative measurements. PET from either scanner modality offered almost the same information despite differences in hardware. Further study is needed to explore features of

  18. Choosing a Scanner: Points To Consider before Buying a Scanner.

    ERIC Educational Resources Information Center

    Raby, Chris

    1998-01-01

    Outlines ten factors to consider before buying a scanner: size of document; type of document; color; speed and volume; resolution; image enhancement; image compression; optical character recognition; scanning subsystem; and the option to use a commercial bureau service. The importance of careful analysis of requirements is emphasized. (AEF)

  19. MR-based Motion Correction for PET Imaging

    PubMed Central

    Ouyang, Jinsong; Li, Quanzheng; Fakhri, Georges El

    2012-01-01

    PET image quality is limited by patient motion. Emission data are blurred due to cardiac and/or respiratory motion. Although spatial resolution is 4 mm for standard clinical whole-body PET scanners, the effective resolution can be a low as 1 cm due to motion. Additionally, the deformation of attenuation medium causes image artifacts. Previously, gating is used to “freeze” the motion, but leads to significantly increased noise level. Simultaneous PET-MR modality offers a new way to perform PET motion correction. MR can be used to measure 3D motion fields, which can then be incorporated into the iterative PET reconstruction to obtain motion corrected PET images. In this report, we present MR imaging techniques to acquire dynamic images, a non-rigid image registration algorithm to extract motion fields from acquired MR images, and a PET reconstruction algorithm with motion correction. We also present results from both phantom and in-vivo animal PET-MR studies. We demonstrate that MR-based PET motion correction using simultaneous PET-MR improves image quality and lesion detectability compared to gating and to no motion correction. PMID:23178089

  20. Evaluation of resistive-plate-chamber-based TOF-PET applied to in-beam particle therapy monitoring

    NASA Astrophysics Data System (ADS)

    Torres-Espallardo, I.; Diblen, F.; Rohling, H.; Solevi, P.; Gillam, J.; Watts, D.; España, S.; Vandenberghe, S.; Fiedler, F.; Rafecas, M.

    2015-05-01

    Particle therapy is a highly conformal radiotherapy technique which reduces the dose deposited to the surrounding normal tissues. In order to fully exploit its advantages, treatment monitoring is necessary to minimize uncertainties related to the dose delivery. Up to now, the only clinically feasible technique for the monitoring of therapeutic irradiation with particle beams is Positron Emission Tomography (PET). In this work we have compared a Resistive Plate Chamber (RPC)-based PET scanner with a scintillation-crystal-based PET scanner for this application. In general, the main advantages of the RPC-PET system are its excellent timing resolution, low cost, and the possibility of building large area systems. We simulated a partial-ring scanner based on an RPC prototype under construction within the Fondazione per Adroterapia Oncologica (TERA). For comparison with the crystal-based PET scanner we have chosen the geometry of a commercially available PET scanner, the Philips Gemini TF. The coincidence time resolution used in the simulations takes into account the current achievable values as well as expected improvements of both technologies. Several scenarios (including patient data) have been simulated to evaluate the performance of different scanners. Initial results have shown that the low sensitivity of the RPC hampers its application to hadron-beam monitoring, which has an intrinsically low positron yield compared to diagnostic PET. In addition, for in-beam PET there is a further data loss due to the partial ring configuration. In order to improve the performance of the RPC-based scanner, an improved version of the RPC detector (modifying the thickness of the gas and glass layers), providing a larger sensitivity, has been simulated and compared with an axially extended version of the crystal-based device. The improved version of the RPC shows better performance than the prototype, but the extended version of the crystal-based PET outperforms all other options.

  1. Implementation and workflow for PET monitoring of therapeutic ion irradiation: a comparison of in-beam, in-room, and off-line techniques

    NASA Astrophysics Data System (ADS)

    Shakirin, Georgy; Braess, Henning; Fiedler, Fine; Kunath, Daniela; Laube, Kristin; Parodi, Katia; Priegnitz, Marlen; Enghardt, Wolfgang

    2011-03-01

    An independent assessment of the dose delivery in ion therapy can be performed using positron emission tomography (PET). For that a distribution of positron emitters which appear as the result of interaction between ions of the therapeutic beam and the irradiated tissue is measured during or after the irradiation. Three concepts for PET monitoring implemented in various therapy facilities are considered in this paper. The in-beam PET concept relies on the PET measurement performed simultaneously to the irradiation by means of a PET scanner which is completely integrated into the irradiation site. The in-room PET concept allows measurement immediately after irradiation by a standalone PET scanner which is installed very close to the irradiation site. In the off-line PET scenario the measurement is performed by means of a standalone PET/CT scanner 10-30 min after the irradiation. These three concepts were evaluated according to image quality criteria, integration costs, and their influence onto the workflow of radiotherapy. In-beam PET showed the best performance. However, the integration costs were estimated as very high for this modality. Moreover, the performance of in-beam PET depends heavily on type and duty cycle of the accelerator. The in-room PET is proposed for planned therapy facilities as a good compromise between the quality of measured data and integration efforts. For facilities which are close to the nuclear medicine departments off-line PET can be suggested under several circumstances.

  2. Lanthanum halide scintillators for time-of-flight 3-D pet

    DOEpatents

    Karp, Joel S.; Surti, Suleman

    2008-06-03

    A Lanthanum Halide scintillator (for example LaCl.sub.3 and LaBr.sub.3) with fast decay time and good timing resolution, as well as high light output and good energy resolution, is used in the design of a PET scanner. The PET scanner includes a cavity for accepting a patient and a plurality of PET detector modules arranged in an approximately cylindrical configuration about the cavity. Each PET detector includes a Lanthanum Halide scintillator having a plurality of Lanthanum Halide crystals, a light guide, and a plurality of photomultiplier tubes arranged respectively peripherally around the cavity. The good timing resolution enables a time-of-flight (TOF) PET scanner to be developed that exhibits a reduction in noise propagation during image reconstruction and a gain in the signal-to-noise ratio. Such a PET scanner includes a time stamp circuit that records the time of receipt of gamma rays by respective PET detectors and provides timing data outputs that are provided to a processor that, in turn, calculates time-of-flight (TOF) of gamma rays through a patient in the cavity and uses the TOF of gamma rays in the reconstruction of images of the patient.

  3. A case study in scanner optimisation

    PubMed Central

    Gibson, NM

    2013-01-01

    Ultrasound scanner preset programmes are factory set or tailored to user requirements. Scanners may, therefore, have different settings for the same application, even on similar equipment in a single department. The aims of this study were: (1) to attempt to match the performance of two scanners, where one was preferred and (2) to assess differences between six scanners used for breast ultrasound within our organisation. The Nottingham Ultrasound Quality Assurance software was used to compare imaging performance. Images of a Gammex RMI 404GS test object were collected from six scanners, using default presets, factory presets and settings matched to a preferred scanner. Resolution, low contrast performance and high contrast performance were measured. The performance of two scanners was successfully matched, where one had been preferred. Default presets varied across the six scanners, three different presets being used. The most used preset differed in settings across the scanners, most notably in the use of different frequency modes. The factory preset was more consistent across the scanners, the main variation being in dynamic range (55–70 dB). Image comparisons showed significant differences, which were reduced or eliminated by adjustment of settings to match a reference scanner. It is possible to match scanner performance using the Nottingham Ultrasound Quality Assurance software as a verification tool. Ultrasound users should be aware that scanners may not behave in a similar fashion, even with apparently equivalent presets. It should be possible to harmonise presets by consensus amongst users.

  4. A Very High Spatial Resolution Detector for Small Animal PET

    SciTech Connect

    Kanai Shah, M.S.

    2007-03-06

    Positron Emission Tomography (PET) is an in vivo analog of autoradiography and has the potential to become a powerful new tool in imaging biological processes in small laboratory animals. PET imaging of small animals can provide unique information that can help in advancement of human disease models as well as drug development. Clinical PET scanners used for human imaging are bulky, expensive and do not have adequate spatial resolution for small animal studies. Hence, dedicated, low cost instruments are required for conducting small animal studies with higher spatial resolution than what is currently achieved with clinical as well as dedicated small animal PET scanners. The goal of the proposed project is to investigate a new all solid-state detector design for small animal PET imaging. Exceptionally high spatial resolution, good timing resolution, and excellent energy resolution are expected from the proposed detector design. The Phase I project was aimed at demonstrating the feasibility of producing high performance solid-state detectors that provide high sensitivity, spatial resolution, and timing characteristics. Energy resolution characteristics of the new detector were also investigated. The goal of the Phase II project is to advance the promising solid-state detector technology for small animal PET and determine its full potential. Detectors modules will be built and characterized and finally, a bench-top small animal PET system will be assembled and evaluated.

  5. Scanner as a Fine Art

    ERIC Educational Resources Information Center

    Fontes, Kris

    2008-01-01

    Not every art department is fortunate enough to have access to digital cameras and image-editing software, but if a scanner, computer, and printer are available, students can create some imaginative and surreal work. This high-school level lesson begins with a discussion of self-portraits, and then moves to students creating images by scanning…

  6. Multispectral scanner (MSS), ERTS-1

    NASA Technical Reports Server (NTRS)

    Arlauskas, J.

    1973-01-01

    The multispectral scanner onboard ERTS-A spacecraft provides simultaneous images in three visible bands and one near infrared band. The instrument employs fiber optics to transfer optical images to the detectors and photomultiplier tubes. Detector outputs are digitized and multiplexed for transmission from the spacecraft by analog to digital processor.

  7. Holographic analyzer and image scanner

    NASA Technical Reports Server (NTRS)

    1978-01-01

    The characteristics, components, and operating procedures are described for a holographic camera real images projection displayer and scanner unit having the capability to upgrade to multiple types of automated raster scan patterns. Schematics of the optical components are included with a diagram of the electric circuit connections.

  8. Pitfalls in PET/CT imaging

    NASA Astrophysics Data System (ADS)

    Rondogianni, Ph; Papathanasiou, N.; Giannopoulou, Ch

    2011-09-01

    PET with 2-[fluorine 18] fluoro-2-deoxy-d-glucose (FDG), has been a clinical tool for the evaluation of various cancers providing valuable metabolic information clinically helpful in the diagnosis, initial staging, therapy monitoring and restaging. However, FDG is not specific for neoplastic processes. Unless anatomic correlation is available to delineate normal structures, pathologic sites of FDG accumulation can easily be confused with normal physiological uptake, leading to false-positive or false-negative findings. Coregistration of PET scans (functional and morphologic information) with computed tomographic (CT) scans (anatomic information) using a combined PET-CT scanner improves the overall sensitivity and specificity of information provided by PET or CT alone. In this paper, we discuss the probable causes of false negative images and pitfalls due to technical reasons, inflammatory processes or benign lesions as well as the utility of PET-CT in differentiating malignant from inflammatory and benign processes, since in some cases such differentiation cannot be made, with certainty, using FDG PET alone.

  9. Imaging quality of (44)Sc in comparison with five other PET radionuclides using Derenzo phantoms and preclinical PET.

    PubMed

    Bunka, Maruta; Müller, Cristina; Vermeulen, Christiaan; Haller, Stephanie; Türler, Andreas; Schibli, Roger; van der Meulen, Nicholas P

    2016-04-01

    PET is the favored nuclear imaging technique because of the high sensitivity and resolution it provides, as well as the possibility for quantification of accumulated radioactivity. (44)Sc (T1/2=3.97h, Eβ(+)=632keV) was recently proposed as a potentially interesting radionuclide for PET. The aim of this study was to investigate the image quality, which can be obtained with (44)Sc, and compare it with five other, frequently employed PET nuclides using Derenzo phantoms and a small-animal PET scanner. The radionuclides were produced at the medical cyclotron at CRS, ETH Zurich ((11)C, (18)F), at the Injector II research cyclotron at CRS, PSI ((64)Cu, (89)Zr, (44)Sc), as well as via a generator system ((68)Ga). Derenzo phantoms, containing solutions of each of these radionuclides, were scanned using a GE Healthcare eXplore VISTA small-animal PET scanner. The image resolution was determined for each nuclide by analysis of the intensity signal using the reconstructed PET data of a hole diameter of 1.3mm. The image quality of (44)Sc was compared to five frequently-used PET radionuclides. In agreement with the positron range, an increasing relative resolution was determined in the sequence of (68)Ga<(44)Sc<(89)Zr<(11)C<(64)Cu<(18)F. The performance of (44)Sc was in agreement with the theoretical expectations based on the energy of the emitted positrons. PMID:26774390

  10. Improvements to Existing Jefferson Lab Wire Scanners

    SciTech Connect

    McCaughan, Michael D.; Tiefenback, Michael G.; Turner, Dennis L.

    2013-06-01

    This poster will detail the augmentation of selected existing CEBAF wire scanners with commercially available hardware, PMTs, and self created software in order to improve the scanners both in function and utility.

  11. Pet Health

    MedlinePlus

    ... Know the signs of medical problems. Take your pet to the veterinarian if you notice: Loss of appetite Drinking a lot of water Gaining or losing a lot of weight quickly Strange behavior Being sluggish and tired Trouble getting up or down Strange lumps

  12. Proceedings of the cardiac PET summit meeting 12 may 2014: Cardiac PET and SPECT instrumentation.

    PubMed

    Garcia, Ernest V

    2015-06-01

    Advances in PET and SPECT and imaging hardware and software are vastly improving the noninvasive evaluation of myocardial perfusion and function. PET perfusion imaging has benefitted from the introduction of novel detectors that now allow true 3D imaging, and precise attenuation correction (AC). These developments have also resulted in perfusion images with higher spatial and contrast resolution that may be acquired in shorter protocols and/or with less patient radiation exposure than traditional PET or SPECT studies. Hybrid PET/CT cameras utilize transmission computed tomographic (CT) scans for AC, and offer the additional clinical advantages of evaluating coronary calcium and myocardial anatomy but at a higher cost than PET scanners that use (68)Ge radioactive line sources. As cardiac PET systems continue to improve, dedicated cardiac SPECT systems are also undergoing a profound change in their design. The scintillation camera general purpose design is being replaced with systems with multiple detectors focused on the heart yielding 5 to 10 times the sensitivity of conventional SPECT. As a result, shorter acquisition times and/or lower tracer doses produce higher quality SPECT images than were possible before. This article reviews these concepts and compares the attributes of PET and SPECT instrumentation. PMID:25824018

  13. A Simple X-Y Scanner.

    ERIC Educational Resources Information Center

    Halse, M. R.; Hudson, W. J.

    1986-01-01

    Describes an X-Y scanner used to create acoustic holograms. Scanner is computer controlled and can be adapted to digitize pictures. Scanner geometry is discussed. An appendix gives equipment details. The control program in ATOM BASIC and 6502 machine code is available from the authors. (JM)

  14. Preliminary results of a prototype C-shaped PET designed for an in-beam PET system

    NASA Astrophysics Data System (ADS)

    Kim, Hyun-Il; Chung, Yong Hyun; Lee, Kisung; Kim, Kyeong Min; Kim, Yongkwon; Joung, Jinhun

    2016-06-01

    Positron emission tomography (PET) can be utilized in particle beam therapy to verify the dose distribution of the target volume as well as the accuracy of the treatment. We present an in-beam PET scanner that can be integrated into a particle beam therapy system. The proposed PET scanner consisted of 14 detector modules arranged in a C-shape to avoid blockage of the particle beam line by the detector modules. Each detector module was composed of a 9×9 array of 4.0 mm×4.0 mm×20.0 mm LYSO crystals optically coupled to four 29-mm-diameter PMTs using the photomultiplier-quadrant-sharing (PQS) technique. In this study, a Geant4 Application for Tomographic Emission (GATE) simulation study was conducted to design a C-shaped PET scanner and then experimental evaluation of the proposed design was performed. The spatial resolution and sensitivity were measured according to NEMA NU2-2007 standards and were 6.1 mm and 5.61 cps/kBq, respectively, which is in good agreement with our simulation, with an error rate of 12.0%. Taken together, our results demonstrate the feasibility of the proposed C-shaped in-beam PET system, which we expect will be useful for measuring dose distribution in particle therapy.

  15. Experimental characterization of the Clear-PEM scanner spectrometric performance

    NASA Astrophysics Data System (ADS)

    Bugalho, R.; Carriço, B.; Ferreira, C. S.; Frade, M.; Ferreira, M.; Moura, R.; Ortigão, C.; Pinheiro, J. F.; Rodrigues, P.; Rolo, I.; Silva, J. C.; Trindade, A.; Varela, J.

    2009-10-01

    In the framework of the Clear-PEM project for the construction of a high-resolution and high-specificity scanner for breast cancer imaging, a Positron Emission Mammography tomograph has been developed and installed at the Instituto Português de Oncologia do Porto hospital. The Clear-PEM scanner is mainly composed by two planar detector heads attached to a robotic arm, trigger/data acquisition electronics system and computing servers. The detector heads hold crystal matrices built from 2 × 2 × 20 mm3 LYSO:Ce crystals readout by Hamamatsu S8550 APD arrays. The APDs are optically coupled to both ends of the 6144 crystals in order to extract the DOI information for each detected event. Each one of 12288 APD's pixels is read and controlled by Application Specific Integrated Circuits water-cooled by an external cooling unit. The Clear-PEM frontend boards innovative design results in a unprecedented integration of the crystal matrices, APDs and ASICs, making Clear-PEM the PET scanner with the highest number of APD pixels ever integrated so far. In this paper, the scanner's main technical characteristics, calibration strategies and the first spectrometric performance evaluation in a clinical environment are presented. The first commissioning results show 99.7% active channels, which, after calibration, have inter-pixel and absolute gain distributions with dispersions of, respectively, 12.2% and 15.3%, demonstrating that despite the large number of channels, the system is uniform. The mean energy resolution at 511 keV is of 15.9%, with a 8.8% dispersion, and the mean CDOI-1 is 5.9%/mm, with a 7.8% dispersion. The coincidence time resolution, at 511 keV, for a energy window between 400 and 600 keV, is 5.2 ns FWHM.

  16. Improving PET imaging for breast cancer using virtual pinhole PET half-ring insert

    NASA Astrophysics Data System (ADS)

    Mathews, Aswin John; Komarov, Sergey; Wu, Heyu; O'Sullivan, Joseph A.; Tai, Yuan-Chuan

    2013-09-01

    A PET insert with detector having smaller crystals and placed near a region of interest in a conventional PET scanner can improve image resolution locally due to the virtual-pinhole PET (VP-PET) effect. This improvement is from the higher spatial sampling of the imaging area near the detector. We have built a prototype half-ring PET insert for head-and-neck cancer imaging applications. In this paper, we extend the use of the insert to breast imaging and show that such a system provides high resolution images of breast and axillary lymph nodes while maintaining the full imaging field of view capability of a clinical PET scanner. We characterize the resolution and contrast recovery for tumors across the imaging field of view. First, we model the system using Monte Carlo methods to determine its theoretical limit of improvement. Simulations were conducted with hot spherical tumors embedded in background activity at tumor-to-background contrast ranging from 3:1 to 12:1. Tumors are arranged in a Derenzo-like pattern with their diameters ranging from 2 to 12 mm. Experimental studies were performed using a chest phantom with cylindrical breast attachment. Tumors of different sizes arranged in a Derenzo-like pattern with tumor-to-background ratio of 6:1 are inserted into the breast phantom. Imaging capability of mediastinum and axillary lymph nodes is explored. Both Monte Carlo simulations and experiment show clear improvement in image resolution and contrast recovery with VP-PET half-ring insert. The degree of improvement in resolution and contrast recovery depends on location of the tumor. The full field of view imaging capability is shown to be maintained. Minor artifacts are introduced in certain regions.

  17. Simultaneous PET and Multispectral 3-Dimensional Fluorescence Optical Tomography Imaging System

    PubMed Central

    Li, Changqing; Yang, Yongfeng; Mitchell, Gregory S.; Cherry, Simon R.

    2015-01-01

    Integrated PET and 3-dimensional (3D) fluorescence optical tomography (FOT) imaging has unique and attractive features for in vivo molecular imaging applications. We have designed, built, and evaluated a simultaneous PET and 3D FOT system. The design of the FOT system is compatible with many existing small-animal PET scanners. Methods The 3D FOT system comprises a novel conical mirror that is used to view the whole-body surface of a mouse with an electron-multiplying charge-coupled device camera when a collimated laser beam is projected on the mouse to stimulate fluorescence. The diffusion equation was used to model the propagation of optical photons inside the mouse body, and 3D fluorescence images were reconstructed iteratively from the fluorescence intensity measurements measured from the surface of the mouse. Insertion of the conical mirror into the gantry of a small-animal PET scanner allowed simultaneous PET and 3D FOT imaging. Results The mutual interactions between PET and 3D FOT were evaluated experimentally. PET has negligible effects on 3D FOT performance. The inserted conical mirror introduces a reduction in the sensitivity and noise-equivalent count rate of the PET system and increases the scatter fraction. PET–FOT phantom experiments were performed. An in vivo experiment using both PET and FOT was also performed. Conclusion Phantom and in vivo experiments demonstrate the feasibility of simultaneous PET and 3D FOT imaging. The first in vivo simultaneous PET–FOT results are reported. PMID:21810591

  18. Optical scanner. [laser doppler velocimeters

    NASA Technical Reports Server (NTRS)

    Rhodes, D. B. (Inventor)

    1977-01-01

    An optical scanner that sequentially focuses optical energy (light) at selected points in space is described. The essential component is a scanning wheel including several glass windows with each window having a different thickness. Due to this difference in thickness, the displacement of the emerging light from the incident light is different for each window. The scanner transmits optical energy to a point in space while at the same time receiving any optical energy generated at that point and then moves on to the next selected point and repeats this transmit and receive operation. It fills the need for a system that permits a laser velocimeter to rapidly scan across a constantly changing flow field in an aerodynamic test facility.

  19. Optimization of PET instrumentation for brain activation studies

    SciTech Connect

    Dahlbom, M.; Cherry, S.R.; Hoffman, E.J. . Dept. of Radiological Science); Eriksson, L. . Dept. of Clinical Neurophysiology); Wienhard, K. )

    1993-08-01

    By performing cerebral blood flow studies with positron emission tomography (PET), and comparing blood flow images of different states of activation, functional mapping of the brain is possible. The ability of current commercial instruments to perform such studies is investigated in this work, based on a comparison of noise equivalent count (NEC) rates. Differences in the NEC performance of the different scanners in conjunction with scanner design parameters, provide insights into the importance of block design (size, dead time, crystal thickness) and overall scanner design (sensitivity and scatter fraction) for optimizing data from activation studies. The newer scanners with removable septa, operating with 3-D acquisition, have much higher sensitivity, but require new methodology for optimized operation. Only by administering multiple low doses (fractionation) of the flow tracer can the high sensitivity be utilized.

  20. Vacuum Attachment for XRF Scanner

    NASA Technical Reports Server (NTRS)

    Schramm, Harry F.; Kaiser, Bruce

    2005-01-01

    Vacuum apparatuses have been developed for increasing the range of elements that can be identified by use of x-ray fluorescent (XRF) scanners of the type mentioned in the two immediately preceding articles. As a consequence of the underlying physical principles, in the presence of air, such an XRF scanner is limited to analysis of chlorine and elements of greater atomic number. When the XRF scanner is operated in a vacuum, it extends the range of analysis to lower atomic numbers - even as far as aluminum and sodium. Hence, more elements will be available for use in XRF labeling of objects as discussed in the two preceding articles. The added benefits of the extended capabilities also have other uses for NASA. Detection of elements of low atomic number is of high interest to the aerospace community. High-strength aluminum alloys will be easily analyzed for composition. Silicon, a major contaminant in certain processes, will be detectable before the process is begun, possibly eliminating weld or adhesion problems. Exotic alloys will be evaluated for composition prior to being placed in service where lives depend on them. And in the less glamorous applications, such as bolts and fasteners, substandard products and counterfeit items will be evaluated at the receiving function and never allowed to enter the operation

  1. IR line scanner on UAV

    NASA Astrophysics Data System (ADS)

    Liu, Shi-chao; Qin, Jie-xin; Qi, Hong-xing; Xiao, Gong-hai

    2011-08-01

    This paper introduces the designing principle and method of the IR line scanner on UAV in three aspects of optical-mechanical system, electronics system and processing software. It makes the system achieve good results in practical application that there are many features in the system such as light weight, small size, low power assumption, wide field of view, high instantaneous field of view, high noise equivalent temperature difference, wirelessly controlled and so on. The entire system is designed as follows: Multi-element scanner is put into use for reducing the electrical noise bandwidth, and then improving SNR; Square split aperture scanner is put into use for solving the image ratation distortion, besides fit for large velocity to height ratio; DSP is put into use for non-uniformity correction and background nosie subtraction, and then improving the imagery quality; SD card is put into use as image data storage media instead of the hard disk; The image data is stored in SD card in FAT32 file system, easily playbacked by processing software on Windows and Linux operating system; wireless transceiver module is put into use for wirelessly controlled.

  2. Robust scanner identification based on noise features

    NASA Astrophysics Data System (ADS)

    Gou, Hongmei; Swaminathan, Ashwin; Wu, Min

    2007-02-01

    A large portion of digital image data available today is acquired using digital cameras or scanners. While cameras allow digital reproduction of natural scenes, scanners are often used to capture hardcopy art in more controlled scenarios. This paper proposes a new technique for non-intrusive scanner model identification, which can be further extended to perform tampering detection on scanned images. Using only scanned image samples that contain arbitrary content, we construct a robust scanner identifier to determine the brand/model of the scanner used to capture each scanned image. The proposed scanner identifier is based on statistical features of scanning noise. We first analyze scanning noise from several angles, including through image de-noising, wavelet analysis, and neighborhood prediction, and then obtain statistical features from each characterization. Experimental results demonstrate that the proposed method can effectively identify the correct scanner brands/models with high accuracy.

  3. Enhanced PET resolution by combining pinhole collimation and coincidence detection.

    PubMed

    DiFilippo, Frank P

    2015-10-21

    Spatial resolution of clinical PET scanners is limited by detector design and photon non-colinearity. Although dedicated small animal PET scanners using specialized high-resolution detectors have been developed, enhancing the spatial resolution of clinical PET scanners is of interest as a more available alternative. Multi-pinhole 511 keV SPECT is capable of high spatial resolution but requires heavily shielded collimators to avoid significant background counts. A practical approach with clinical PET detectors is to combine multi-pinhole collimation with coincidence detection. In this new hybrid modality, there are three locations associated with each event, namely those of the two detected photons and the pinhole aperture. These three locations over-determine the line of response and provide redundant information that is superior to coincidence detection or pinhole collimation alone. Multi-pinhole collimation provides high resolution and avoids non-colinearity error but is subject to collimator penetration and artifacts from overlapping projections. However the coincidence information, though at lower resolution, is valuable for determining whether the photon passed near a pinhole within the cone acceptance angle and for identifying through which pinhole the photon passed. This information allows most photons penetrating through the collimator to be rejected and avoids overlapping projections. With much improved event rejection, a collimator with minimal shielding may be used, and a lightweight add-on collimator for high resolution imaging is feasible for use with a clinical PET scanner. Monte Carlo simulations were performed of a (18)F hot rods phantom and a 54-pinhole unfocused whole-body mouse collimator with a clinical PET scanner. Based on coincidence information and pinhole geometry, events were accepted or rejected, and pinhole-specific crystal-map projections were generated. Tomographic images then were reconstructed using a conventional pinhole SPECT

  4. Enhanced PET resolution by combining pinhole collimation and coincidence detection

    NASA Astrophysics Data System (ADS)

    DiFilippo, Frank P.

    2015-10-01

    Spatial resolution of clinical PET scanners is limited by detector design and photon non-colinearity. Although dedicated small animal PET scanners using specialized high-resolution detectors have been developed, enhancing the spatial resolution of clinical PET scanners is of interest as a more available alternative. Multi-pinhole 511 keV SPECT is capable of high spatial resolution but requires heavily shielded collimators to avoid significant background counts. A practical approach with clinical PET detectors is to combine multi-pinhole collimation with coincidence detection. In this new hybrid modality, there are three locations associated with each event, namely those of the two detected photons and the pinhole aperture. These three locations over-determine the line of response and provide redundant information that is superior to coincidence detection or pinhole collimation alone. Multi-pinhole collimation provides high resolution and avoids non-colinearity error but is subject to collimator penetration and artifacts from overlapping projections. However the coincidence information, though at lower resolution, is valuable for determining whether the photon passed near a pinhole within the cone acceptance angle and for identifying through which pinhole the photon passed. This information allows most photons penetrating through the collimator to be rejected and avoids overlapping projections. With much improved event rejection, a collimator with minimal shielding may be used, and a lightweight add-on collimator for high resolution imaging is feasible for use with a clinical PET scanner. Monte Carlo simulations were performed of a 18F hot rods phantom and a 54-pinhole unfocused whole-body mouse collimator with a clinical PET scanner. Based on coincidence information and pinhole geometry, events were accepted or rejected, and pinhole-specific crystal-map projections were generated. Tomographic images then were reconstructed using a conventional pinhole SPECT

  5. Incorporating anatomical side information into PET reconstruction using nonlocal regularization.

    PubMed

    Nguyen, Van-Giang; Lee, Soo-Jin

    2013-10-01

    With the introduction of combined positron emission tomography (PET)/computed tomography (CT) or PET/magnetic resonance imaging (MRI) scanners, there is an increasing emphasis on reconstructing PET images with the aid of the anatomical side information obtained from X-ray CT or MRI scanners. In this paper, we propose a new approach to incorporating prior anatomical information into PET reconstruction using the nonlocal regularization method. The nonlocal regularizer developed for this application is designed to selectively consider the anatomical information only when it is reliable. As our proposed nonlocal regularization method does not directly use anatomical edges or boundaries which are often used in conventional methods, it is not only free from additional processes to extract anatomical boundaries or segmented regions, but also more robust to the signal mismatch problem that is caused by the indirect relationship between the PET image and the anatomical image. We perform simulations with digital phantoms. According to our experimental results, compared to the conventional method based on the traditional local regularization method, our nonlocal regularization method performs well even with the imperfect prior anatomical information or in the presence of signal mismatch between the PET image and the anatomical image. PMID:23744678

  6. Clinical Application of in-room PET for in vivo Treatment Monitoring in Proton Radiotherapy

    PubMed Central

    Min, Chul Hee; Zhu, Xuping; Winey, Brian A.; Grogg, Kira; Testa, Mauro; Fakhri, Georges El; Bortfeld, Thomas R.; Paganetti, Harald; Shih, Helen A.

    2013-01-01

    Purpose/Objective(s) The purpose of this study is to evaluate the potential of using an in-room PET for treatment verification in proton therapy and to derive suitable PET scan times. Materials/Methods Nine patients undergoing passive scattering proton therapy were scanned immediately after treatment with an in-room PET scanner. The scanner was positioned next to the treatment head after treatment. The Monte Carlo (MC) method was employed to reproduce PET activities for each patient. To assess the proton beam range uncertainty we designed a novel concept where the measured PET activity surface distal to the target at the end of range was compared with MC predictions. The repositioning of patients for the PET scan took on average about 2 minutes. The PET images were reconstructed considering varying scan times to test the scan time dependency of the method. Results The measured PET images show overall good spatial correlations with MC predictions. Some discrepancies could be attributed to uncertainties in the local elemental composition and biological washout. For 8 patients treated with a single field, the average range differences between PET measurements and CT-image-based MC results were less than 5 mm (< 3 mm for 6 of 8 patients) and root-mean-square deviations (RMSD) were 4-11 mm with PET-CT image co-registration errors of about 2 mm. Our results also show that a short-length PET scan of 5 minutes can yield similar results compared to a 20 minutes PET scan. Conclusions Our first clinical trials of 9 patients using an in-room PET system demonstrated its potential for in vivo treatment monitoring in proton therapy. For a quantitative range prediction with arbitrary shape of target volume, we suggest employing the distal PET activity surface. PMID:23391817

  7. Novel Developments in Instrumentation for PET Imaging

    NASA Astrophysics Data System (ADS)

    Karp, Joel

    2013-04-01

    Advances in medical imaging, in particular positron emission tomography (PET), have been based on technical developments in physics and instrumentation that have common foundations with detection systems used in other fields of physics. New detector materials are used in PET systems that maximize efficiency, timing characteristics and robustness, and which lead to improved image quality and quantitative accuracy for clinical imaging. Time of flight (TOF) techniques are now routinely used in commercial PET scanners that combine physiological imaging with anatomical imaging provided by x-ray computed tomography. Using new solid-state photo-sensors instead of traditional photo-multiplier tubes makes it possible to combine PET with magnetic resonance imaging which is a significant technical challenge, but one that is creating new opportunities for both research and clinical applications. An overview of recent advances in instrumentation, such as TOF and PET/MR will be presented, along with examples of imaging studies to demonstrate the impact on patient care and basic research of diseases.

  8. A large area, silicon photomultiplier-based PET detector module

    PubMed Central

    Raylman, RR; Stolin, A; Majewski, S; Proffitt, J

    2013-01-01

    The introduction of silicon photomultipliers (SiPM) has facilitated construction of compact, efficient and magnetic field-hardened positron emission tomography (PET) scanners. To take full advantage of these devices, methods for using them to produce large field-of-view PET scanners are needed. In this investigation, we explored techniques to combine two SiPM arrays to form the building block for a small animal PET scanner. The module consists of a 26 × 58 array of 1.5 × 1.5mm2 LYSO elements (spanning 41 × 91mm2) coupled to two SensL SiPM arrays. The SiPMs were read out with new multiplexing electronics developed for this project. To facilitate calculation of event position with multiple SiPM arrays it was necessary to spread scintillation light amongst a number of elements with a small light guide. This method was successful in permitting identification of all detector elements, even at the seam between two SiPM arrays. Since the performance of SiPMs is enhanced by cooling, the detector module was fitted with a cooling jacket, which allowed the temperature of the device and electronics to be controlled. Testing demonstrated that the peak-to-valley contrast ratio of the light detected from the scintillation array was increased by ∼45% when the temperature was reduced from 28 °C to 16 °C. Energy resolution for 511 keV photons improved slightly from 18.8% at 28 °C to 17.8% at 16 °C. Finally, the coincidence timing resolution of the module was found to be insufficient for time-of-flight applications (∼2100 ps at 14 °C). The first use of these new modules will be in the construction of a small animal PET scanner to be integrated with a 3T clinical magnetic resonance imaging scanner. PMID:24319305

  9. A large area, silicon photomultiplier-based PET detector module

    NASA Astrophysics Data System (ADS)

    Raylman, R. R.; Stolin, A.; Majewski, S.; Proffitt, J.

    2014-01-01

    The introduction of silicon photomultipliers (SiPM) has facilitated construction of compact, efficient and magnetic field-hardened positron emission tomography (PET) scanners. To take full advantage of these devices, methods for using them to produce large field-of-view PET scanners are needed. In this investigation, we explored techniques to combine two SiPM arrays to form the building block for a small animal PET scanner. The module consists of a 26×58 array of 1.5×1.5 mm2 LYSO elements (spanning 41×91 mm2) coupled to two SensL SiPM arrays. The SiPMs were read out with new multiplexing electronics developed for this project. To facilitate calculation of event position with multiple SiPM arrays it was necessary to spread scintillation light amongst a number of elements with a small light guide. This method was successful in permitting identification of all detector elements, even at the seam between two SiPM arrays. Since the performance of SiPMs is enhanced by cooling, the detector module was fitted with a cooling jacket, which allowed the temperature of the device and electronics to be controlled. Testing demonstrated that the peak-to-valley contrast ratio of the light detected from the scintillation array was increased by ~45% when the temperature was reduced from 28 °C to 16 °C. Energy resolution for 511 keV photons improved slightly from 18.8% at 28 °C to 17.8% at 16 °C. Finally, the coincidence timing resolution of the module was found to be insufficient for time-of-flight applications (~2100 ps at 14 °C). The first use of these new modules will be in the construction of a small animal PET scanner to be integrated with a 3 T clinical magnetic resonance imaging scanner.

  10. Pet Problems at Home: Pet Problems in the Community.

    ERIC Educational Resources Information Center

    Soltow, Willow

    1984-01-01

    Discusses problems of pets in the community, examining the community's role related to disruptive pets and pet overpopulation. Also discusses pet problems at home, offering advice on selecting a pet, meeting a pet's needs, and disciplining pets. Includes a list of books, films/filmstrips, teaching materials, and various instructional strategies.…

  11. Influence of Iterative Reconstruction Algorithms on PET Image Resolution

    NASA Astrophysics Data System (ADS)

    Karpetas, G. E.; Michail, C. M.; Fountos, G. P.; Valais, I. G.; Nikolopoulos, D.; Kandarakis, I. S.; Panayiotakis, G. S.

    2015-09-01

    The aim of the present study was to assess image quality of PET scanners through a thin layer chromatography (TLC) plane source. The source was simulated using a previously validated Monte Carlo model. The model was developed by using the GATE MC package and reconstructed images obtained with the STIR software for tomographic image reconstruction. The simulated PET scanner was the GE DiscoveryST. A plane source consisted of a TLC plate, was simulated by a layer of silica gel on aluminum (Al) foil substrates, immersed in 18F-FDG bath solution (1MBq). Image quality was assessed in terms of the modulation transfer function (MTF). MTF curves were estimated from transverse reconstructed images of the plane source. Images were reconstructed by the maximum likelihood estimation (MLE)-OSMAPOSL, the ordered subsets separable paraboloidal surrogate (OSSPS), the median root prior (MRP) and OSMAPOSL with quadratic prior, algorithms. OSMAPOSL reconstruction was assessed by using fixed subsets and various iterations, as well as by using various beta (hyper) parameter values. MTF values were found to increase with increasing iterations. MTF also improves by using lower beta values. The simulated PET evaluation method, based on the TLC plane source, can be useful in the resolution assessment of PET scanners.

  12. Pet Bonding and Pet Bereavement among Adolescents.

    ERIC Educational Resources Information Center

    Brown, Brenda H.; And Others

    1996-01-01

    Studied adolescent-pet bonding and bereavement following pet loss (n=55). Hypothesized that highly-bonded adolescents experience more intense grief when a pet dies than do those less bonded; degree of bonding is greater for girls than for boys; and intensity of bereavement is greater for girls than for boys. Results supported the hypotheses. (RB)

  13. 18F-FDG PET/CT and PET/MRI Perform Equally Well in Cancer: Evidence from Studies on More Than 2,300 Patients.

    PubMed

    Spick, Claudio; Herrmann, Ken; Czernin, Johannes

    2016-03-01

    (18)F-FDG PET/CT has become the reference standard in oncologic imaging against which the performance of other imaging modalities is measured. The promise of PET/MRI includes multiparametric imaging to further improve diagnosis and phenotyping of cancer. Rather than focusing on these capabilities, many investigators have examined whether (18)F-FDG PET combined with mostly anatomic MRI improves cancer staging and restaging. After a description of PET/MRI scanner designs and a discussion of technical and operational issues, we review the available literature to determine whether cancer assessments are improved with PET/MRI. The available data show that PET/MRI is feasible and performs as well as PET/CT in most types of cancer. Diagnostic advantages may be achievable in prostate cancer and in bone metastases, whereas disadvantages exist in lung nodule assessments. We conclude that (18)F-FDG PET/MRI and PET/CT provide comparable diagnostic information when MRI is used simply to provide the anatomic framework. Thus, PET/MRI could be used in lieu of PET/CT if this approach becomes economically viable and if reasonable workflows can be established. Future studies should explore the multiparametric potential of MRI. PMID:26742709

  14. PET/MR Imaging in Cancers of the Gastrointestinal Tract.

    PubMed

    Paspulati, Raj Mohan; Gupta, Amit

    2016-10-01

    PET/computed tomography (PET/CT) is an established hybrid imaging technique for staging and follow-up of gastrointestinal (GI) tract malignancies, especially for colorectal carcinoma. Dedicated hybrid PET/MR imaging scanners are currently available for clinical use. Although they will not replace regular use of PET/CT, they may have utility in selected cases of GI tract malignancies. The superior soft tissue contrast resolution and depiction of anatomy and the functional information obtained from diffusion-weighted imaging (DWI) provided by MR imaging in PET/MR imaging are advantages over CT of PET/CT for T staging and follow-up of rectal carcinoma and for better characterization of liver lesions. Functional information from DWI and use of liver-specific MR imaging contrast agents are an added advantage in follow-up of liver metastases after systemic and locoregional treatment. New radiotracers will improve the utility of PET/MR imaging in staging and follow-up of tumors, which may not be [18F]-2-fluoro-2-deoxy-d-glucose avid, such as hepatocellular carcinoma and neuroendocrine tumors. PET/MR imaging also has application in selected cases of cholangiocarcinoma, gallbladder cancer, and pancreatic carcinoma for initial staging and follow-up assessment. PMID:27593246

  15. The current state, challenges and perspectives of MR-PET.

    PubMed

    Herzog, Hans; Pietrzyk, Uwe; Shah, N Jon; Ziemons, Karl

    2010-02-01

    Following the success of PET/CT during the last decade and the recent increasing proliferation of SPECT/CT, another hybrid imaging instrument has been gaining more and more interest: MR-PET. First combined, simultaneous PET and MR studies carried out in small animals demonstrated the feasibility of the new approach. Concurrently, some prototypes of an MR-PET scanner for simultaneous human brain studies have been built, their performance is being tested and preliminary applications have already been shown. Through this pioneering work, it has become clear that advances in the detector design are necessary for further optimization. Recently, the different issues related to the present state and future prospects of MR-PET were presented and discussed during an international 2-day workshop at the Forschungszentrum Jülich, Germany, held after, and in conjunction with, the 2008 IEEE Nuclear Science Symposium and Medical Imaging Conference in Dresden, Germany on October 27-28, 2008. The topics ranged from small animal MR-PET imaging to human MR-BrainPET imaging, new detector developments, challenges/opportunities for ultra-high field MR-PET imaging and considerations of possible future research and clinical applications. This report presents a critical summary of the contributions made to the workshop. PMID:19853045

  16. Quantitative evaluation of PET image using event information bootstrap

    NASA Astrophysics Data System (ADS)

    Song, Hankyeol; Kwak, Shin Hye; Kim, Kyeong Min; Kang, Joo Hyun; Chung, Yong Hyun; Woo, Sang-Keun

    2016-04-01

    The purpose of this study was to enhance the effect in the PET image quality according to event bootstrap of small animal PET data. In order to investigate the time difference condition, realigned sinograms were generated from randomly sampled data set using bootstrap. List-mode data was obtained from small animal PET scanner for Ge-68 30 sec, Y-90 20 min and Y-90 60 min. PET image was reconstructed by Ordered Subset Expectation Maximization(OSEM) 2D with the list-mode format. Image analysis was investigated by Signal to Noise Ratio(SNR) of Ge-68 and Y-90 image. Non-parametric resampled PET image SNR percent change for the Ge-68 30 sec, Y-90 60 min, and Y-90 20 min was 1.69 %, 7.03 %, and 4.78 %, respectively. SNR percent change of non-parametric resampled PET image with time difference condition was 1.08 % for the Ge-68 30 sec, 6.74 % for the Y-90 60 min and 10.94 % for the Y-90 29 min. The result indicated that the bootstrap with time difference condition had a potential to improve a noisy Y-90 PET image quality. This method should be expected to reduce Y-90 PET measurement time and to enhance its accuracy.

  17. Recovery and normalization of triple coincidences in PET

    SciTech Connect

    Lage, Eduardo Parot, Vicente; Dave, Shivang R.; Herraiz, Joaquin L.; Moore, Stephen C.; Sitek, Arkadiusz; Park, Mi-Ae; Udías, Jose M.; Vaquero, Juan J.

    2015-03-15

    Purpose: Triple coincidences in positron emission tomography (PET) are events in which three γ-rays are detected simultaneously. These events, though potentially useful for enhancing the sensitivity of PET scanners, are discarded or processed without special consideration in current systems, because there is not a clear criterion for assigning them to a unique line-of-response (LOR). Methods proposed for recovering such events usually rely on the use of highly specialized detection systems, hampering general adoption, and/or are based on Compton-scatter kinematics and, consequently, are limited in accuracy by the energy resolution of standard PET detectors. In this work, the authors propose a simple and general solution for recovering triple coincidences, which does not require specialized detectors or additional energy resolution requirements. Methods: To recover triple coincidences, the authors’ method distributes such events among their possible LORs using the relative proportions of double coincidences in these LORs. The authors show analytically that this assignment scheme represents the maximum-likelihood solution for the triple-coincidence distribution problem. The PET component of a preclinical PET/CT scanner was adapted to enable the acquisition and processing of triple coincidences. Since the efficiencies for detecting double and triple events were found to be different throughout the scanner field-of-view, a normalization procedure specific for triple coincidences was also developed. The effect of including triple coincidences using their method was compared against the cases of equally weighting the triples among their possible LORs and discarding all the triple events. The authors used as figures of merit for this comparison sensitivity, noise-equivalent count (NEC) rates and image quality calculated as described in the NEMA NU-4 protocol for the assessment of preclinical PET scanners. Results: The addition of triple-coincidence events with the

  18. 18F-FDG PET of the hands with a dedicated high-resolution PEM system (arthro-PET): correlation with PET/CT, radiography and clinical parameters

    PubMed Central

    Mhlanga, Joyce C.; Carrino, John A.; Lodge, Martin; Wang, Hao

    2015-01-01

    Purpose The aim of this study was to prospectively determine the feasibility and compare the novel use of a positron emission mammography (PEM) scanner with standard PET/CT for evaluating hand osteoarthritis (OA) with 18F-FDG. Methods Institutional review board approval and written informed consent were obtained for this HIPAA-compliant prospective study in which 14 adults referred for oncological 18F-FDG PET/CT underwent dedicated hand PET/CT followed by arthro-PET using the PEM device. Hand radiographs were obtained and scored for the presence and severity of OA. Summed qualitative and quantitative joint glycolytic scores for each modality were compared with the findings on plain radiography and clinical features. Results Eight patients with clinical and/or radiographic evidence of OA comprised the OA group (mean age 73±7.7 years). Six patients served as the control group (53.7±9.3 years). Arthro-PET quantitative and qualitative joint glycolytic scores were highly correlated with PET/CT findings in the OA patients (r=0.86. p =0.007; r=0.94, p=0.001). Qualitative arthro-PET and PET/CT joint scores were significantly higher in the OA patients than in controls (38.7±6.6 vs. 32.2±0.4, p=0.02; 37.5±5.4 vs. 32.2±0.4, p=0.03, respectively). Quantitative arthro-PET and PET/CT maximum SUV-lean joint scores were higher in the OA patients, although they did not reach statistical significance (20.8±4.2 vs. 18±1.8, p= 0.13; 22.8±5.38 vs. 20.1±1.54, p=0.21). By definition, OA patients had higher radiographic joint scores than controls (30.9±31.3 vs. 0, p=0.03). Conclusion Hand imaging using a small field of view PEM system (arthro-PET) with FDG is feasible, performing comparably to PET/CT in assessing metabolic joint activity. Arthro-PET and PET/CT showed higher joint FDG uptake in OA. Further exploration of arthro-PET in arthritis management is warranted. PMID:25134669

  19. Development of a novel linearly-filled Derenzo microPET phantom.

    PubMed

    Cox, Benjamin L; Graves, Stephen A; Farhoud, Mohammed; Barnhart, Todd E; Jeffery, Justin J; Eliceiri, Kevin W; Nickles, Robert J

    2016-01-01

    Positron emission tomography (PET) phantoms are used to calibrate PET scanners so that inter-scanner and inter-isotope comparison can be made between PET datasets. Hot rod style phantoms have a hole pattern, which is filled with a positron-emitting isotope and typically involves using two radioisotope reservoirs with the pattern created with channels in between. However, this configuration is difficult to fill and requires an excess of activity and volume. Here we present an alternative design, a phantom that is linearly filled-one channel at a time. The process of fabrication of prototypes of the design is described and PET images of the prototyped phantom are also shown for a variety of commonly used radioisotopes ((52)Mn, (64)Cu, (76)Br, (124)I). This design allows for a large reduction in isotope volume and required filling time making a quality assurance (QA) protocol safer, more efficient and less costly. PMID:27508106

  20. Development of a novel linearly-filled Derenzo microPET phantom

    PubMed Central

    Cox, Benjamin L; Graves, Stephen A; Farhoud, Mohammed; Barnhart, Todd E; Jeffery, Justin J; Eliceiri, Kevin W; Nickles, Robert J

    2016-01-01

    Positron emission tomography (PET) phantoms are used to calibrate PET scanners so that inter-scanner and inter-isotope comparison can be made between PET datasets. Hot rod style phantoms have a hole pattern, which is filled with a positron-emitting isotope and typically involves using two radioisotope reservoirs with the pattern created with channels in between. However, this configuration is difficult to fill and requires an excess of activity and volume. Here we present an alternative design, a phantom that is linearly filled-one channel at a time. The process of fabrication of prototypes of the design is described and PET images of the prototyped phantom are also shown for a variety of commonly used radioisotopes (52Mn, 64Cu, 76Br, 124I). This design allows for a large reduction in isotope volume and required filling time making a quality assurance (QA) protocol safer, more efficient and less costly. PMID:27508106

  1. Impact of MR based attenuation correction on neurological PET studies

    PubMed Central

    Su, Yi; Rubin, Brian B.; McConathy, Jonathan; Laforest, Richard; Qi, Jing; Sharma, Akash; Priatna, Agus; Benzinger, Tammie L.S.

    2016-01-01

    Hybrid positron emission tomography (PET) and magnetic resonance (MR) scanners have become a reality in recent years with the benefits of reduced radiation exposure, reduction of imaging time, and potential advantages in quantification. Appropriate attenuation correction remains a challenge. Biases in PET activity measurements were demonstrated using the current MR based attenuation correction technique. We aim to investigate the impact of using standard MRAC technique on the clinical and research utility of PET/MR hybrid scanner for amyloid imaging. Methods Florbetapir scans were obtained on 40 participants on a Biograph mMR hybrid scanner with simultaneous MR acquisition. PET images were reconstructed using both MR and CT derived attenuation map. Quantitative analysis was performed for both datasets to assess the impact of MR based attenuation correction to absolute PET activity measurements as well as target to reference ratio (SUVR). Clinical assessment was also performed by a nuclear medicine physician to determine amyloid status based on the criteria in the FDA prescribing information for florbetapir. Results MR based attenuation correction led to underestimation of PET activity for most part of the brain with a small overestimation for deep brain regions. There is also an overestimation of SUVR values with cerebellar reference. SUVR measurements obtained from the two attenuation correction methods were strongly correlated. Clinical assessment of amyloid status resulted in identical classification as positive or negative regardless of the attenuation correction methods. Conclusions MR based attenuation correction cause biases in quantitative measurements. The biases may be accounted for by a linear model, although the spatial variation cannot be easily modelled. The quantitative differences however did not affect clinical assessment as positive or negative. PMID:26823562

  2. Small animal simultaneous PET/MRI: initial experiences in a 9.4T microMRI

    SciTech Connect

    Maramraju, S.H.; Schlyer, D.; Maramraju, S.H.; Smith, S.D.; Junnarkar, S.S.; Schulz, D.; Stoll, S.; Ravindranath, B.; Purschke, M.L.; Rescia, S.; Southekal, S.; Pratte, J.-F.; Vaska, P.; Woody, C.L.; Schlyer, D.J.

    2011-03-25

    We developed a non-magnetic positron-emission tomography (PET) device based on the rat conscious animal PET that operates in a small-animal magnetic resonance imaging (MRI) scanner, thereby enabling us to carry out simultaneous PET/MRI studies. The PET detector comprises 12 detector blocks, each being a 4 x 8 array of lutetium oxyorthosilicate crystals (2.22 x 2.22 x 5 mm{sup 3}) coupled to a matching non-magnetic avalanche photodiode array. The detector blocks, housed in a plastic case, form a 38 mm inner diameter ring with an 18 mm axial extent. Custom-built MRI coils fit inside the positron-emission tomography (PET) device, operating in transceiver mode. The PET insert is integrated with a Bruker 9.4 T 210 mm clear-bore diameter MRI scanner. We acquired simultaneous PET/MR images of phantoms, of in vivo rat brain, and of cardiac-gated mouse heart using [{sup 11}C]raclopride and 2-deoxy-2-[{sup 18}F]fluoro-d-glucose PET radiotracers. There was minor interference between the PET electronics and the MRI during simultaneous operation, and small effects on the signal-to-noise ratio in the MR images in the presence of the PET, but no noticeable visual artifacts. Gradient echo and high-duty-cycle spin echo radio frequency (RF) pulses resulted in a 7% and a 28% loss in PET counts, respectively, due to high PET counts during the RF pulses that had to be gated out. The calibration of the activity concentration of PET data during MR pulsing is reproducible within less than 6%. Our initial results demonstrate the feasibility of performing simultaneous PET and MRI studies in adult rats and mice using the same PET insert in a small-bore 9.4 T MRI.

  3. Small animal simultaneous PET/MRI: initial experiences in a 9.4 T microMRI.

    PubMed

    Maramraju, Sri Harsha; Smith, S David; Junnarkar, Sachin S; Schulz, Daniela; Stoll, Sean; Ravindranath, Bosky; Purschke, Martin L; Rescia, Sergio; Southekal, Sudeepti; Pratte, Jean-François; Vaska, Paul; Woody, Craig L; Schlyer, David J

    2011-04-21

    We developed a non-magnetic positron-emission tomography (PET) device based on the rat conscious animal PET that operates in a small-animal magnetic resonance imaging (MRI) scanner, thereby enabling us to carry out simultaneous PET/MRI studies. The PET detector comprises 12 detector blocks, each being a 4 × 8 array of lutetium oxyorthosilicate crystals (2.22 × 2.22 × 5 mm(3)) coupled to a matching non-magnetic avalanche photodiode array. The detector blocks, housed in a plastic case, form a 38 mm inner diameter ring with an 18 mm axial extent. Custom-built MRI coils fit inside the positron-emission tomography (PET) device, operating in transceiver mode. The PET insert is integrated with a Bruker 9.4 T 210 mm clear-bore diameter MRI scanner. We acquired simultaneous PET/MR images of phantoms, of in vivo rat brain, and of cardiac-gated mouse heart using [(11)C]raclopride and 2-deoxy-2-[(18)F]fluoro-D-glucose PET radiotracers. There was minor interference between the PET electronics and the MRI during simultaneous operation, and small effects on the signal-to-noise ratio in the MR images in the presence of the PET, but no noticeable visual artifacts. Gradient echo and high-duty-cycle spin echo radio frequency (RF) pulses resulted in a 7% and a 28% loss in PET counts, respectively, due to high PET counts during the RF pulses that had to be gated out. The calibration of the activity concentration of PET data during MR pulsing is reproducible within less than 6%. Our initial results demonstrate the feasibility of performing simultaneous PET and MRI studies in adult rats and mice using the same PET insert in a small-bore 9.4 T MRI. PMID:21441651

  4. Small animal simultaneous PET/MRI: initial experiences in a 9.4 T microMRI

    NASA Astrophysics Data System (ADS)

    Harsha Maramraju, Sri; Smith, S. David; Junnarkar, Sachin S.; Schulz, Daniela; Stoll, Sean; Ravindranath, Bosky; Purschke, Martin L.; Rescia, Sergio; Southekal, Sudeepti; Pratte, Jean-François; Vaska, Paul; Woody, Craig L.; Schlyer, David J.

    2011-04-01

    We developed a non-magnetic positron-emission tomography (PET) device based on the rat conscious animal PET that operates in a small-animal magnetic resonance imaging (MRI) scanner, thereby enabling us to carry out simultaneous PET/MRI studies. The PET detector comprises 12 detector blocks, each being a 4 × 8 array of lutetium oxyorthosilicate crystals (2.22 × 2.22 × 5 mm3) coupled to a matching non-magnetic avalanche photodiode array. The detector blocks, housed in a plastic case, form a 38 mm inner diameter ring with an 18 mm axial extent. Custom-built MRI coils fit inside the positron-emission tomography (PET) device, operating in transceiver mode. The PET insert is integrated with a Bruker 9.4 T 210 mm clear-bore diameter MRI scanner. We acquired simultaneous PET/MR images of phantoms, of in vivo rat brain, and of cardiac-gated mouse heart using [11C]raclopride and 2-deoxy-2-[18F]fluoro-d-glucose PET radiotracers. There was minor interference between the PET electronics and the MRI during simultaneous operation, and small effects on the signal-to-noise ratio in the MR images in the presence of the PET, but no noticeable visual artifacts. Gradient echo and high-duty-cycle spin echo radio frequency (RF) pulses resulted in a 7% and a 28% loss in PET counts, respectively, due to high PET counts during the RF pulses that had to be gated out. The calibration of the activity concentration of PET data during MR pulsing is reproducible within less than 6%. Our initial results demonstrate the feasibility of performing simultaneous PET and MRI studies in adult rats and mice using the same PET insert in a small-bore 9.4 T MRI.

  5. Design and evaluation of the MAMMI dedicated breast PET

    SciTech Connect

    Moliner, L.; Gonzalez, A. J.; Soriano, A.; Sanchez, F.; Correcher, C.; Orero, A.; Carles, M.; Vidal, L. F.; Barbera, J.; Caballero, L.; Seimetz, M.; Vazquez, C.; Benlloch, J. M.

    2012-09-15

    Purpose: A breast dedicated positron emission tomography (PET) scanner has been developed based on monolithic LYSO crystals coupled to position sensitive photomultiplier tubes (PSPMTs). In this study, we describe the design of the PET system and report on its performance evaluation. Methods: MAMMI is a breast PET scanner based on monolithic LYSO crystals. It consists of 12 compact modules with a transaxial field of view (FOV) of 170 mm in diameter and 40 mm axial FOV that translates to cover up to 170 mm. The patient lies down in a prone position that facilitates maximum breast elongation. Quantitative performance analysis of the calculated method for the attenuation correction specifically developed for MAMMI, and based on PET image segmentation, has also been conducted in this evaluation. In order to fully determine the MAMMI prototype's performance, we have adapted the measurements suggested for National Electrical Manufacturers Association (NEMA) NU 2-2007 and NU 4-2008 protocol tests, as they are defined for whole-body and small animal PET scanners, respectively. Results: Spatial resolutions of 1.6, 1.8, and 1.9 mm were measured in the axial, radial, and tangential directions, respectively. A scatter fraction of 20.8% was obtained and the maximum NEC was determined to be 25 kcps at 44 MBq. The average sensitivity of the system was observed to be 1% for an energy window of (250 keV-750 keV) and a maximum absolute sensitivity of 1.8% was measured at the FOV center. Conclusions: The overall performance of the MAMMI reported on this evaluation quantifies its ability to produce high quality PET images. Spatial resolution values below 3 mm were measured in most of the FOV. Only the radial component of spatial resolution exceeds the 3 mm at radial positions larger than 60 mm. This study emphasizes the need for standardized testing methodologies for dedicated breast PET systems similar to NEMA standards for whole-body and small animal PET scanners.

  6. X-ray microtomographic scanners

    SciTech Connect

    Syryamkin, V. I. Klestov, S. A.

    2015-11-17

    The article studies the operating procedures of an X-ray microtomographic scanner and the module of reconstruction and analysis 3D-image of a test sample in particular. An algorithm for 3D-image reconstruction based on image shadow projections and mathematical methods of the processing are described. Chapter 1 describes the basic principles of X-ray tomography and general procedures of the device developed. Chapters 2 and 3 are devoted to the problem of resources saving by the system during the X-ray tomography procedure, which is achieved by preprocessing of the initial shadow projections. Preprocessing includes background noise removing from the images, which reduces the amount of shadow projections in general and increases the efficiency of the group shadow projections compression. In conclusion, the main applications of X-ray tomography are presented.

  7. Non-Destructive Testing Scanner

    NASA Technical Reports Server (NTRS)

    1990-01-01

    Bio-Imaging Research's technology that originated in an aerospace program has come full circle with a new aerospace adaptation called the Advanced Computed Tomography Inspection System, or ACTIS. The medical version of CT scans the human body for tumors or other abnormalities, the ACTIS system finds imperfections in aerospace structures and components, such as castings, assemblies, rocket motors and nozzles. ACTIS is described by its developer as the most versatile CT scanner available for non-destructive testing applications. ACTIS is a variable geometry system. ACTIS source and detectors can be moved closer together or farther apart to optimize the geometry for different sizes of test objects. The combination of variable geometry, three sources, and focusing detectors makes ACTIS cost effective for a broad range of applications. System can scan anything from very small turbine blades to large rocket assemblies.

  8. Code-multiplexed optical scanner

    NASA Astrophysics Data System (ADS)

    Riza, Nabeel A.; Arain, Muzammil A.

    2003-03-01

    A three-dimensional (3-D) optical-scanning technique is proposed based on spatial optical phase code activation on an input beam. This code-multiplexed optical scanner (C-MOS) relies on holographically stored 3-D beam-forming information. Proof-of-concept C-MOS experimental results by use of a photorefractive crystal as a holographic medium generates eight beams representing a basic 3-D voxel element generated via a binary-code matrix of the Hadamard type. The experiment demonstrates the C-MOS features of no moving parts, beam-forming flexibility, and large centimeter-size apertures. A novel application of the C-MOS as an optical security lock is highlighted.

  9. Laser Scanner For Automatic Storage

    NASA Astrophysics Data System (ADS)

    Carvalho, Fernando D.; Correia, Bento A.; Rebordao, Jose M.; Rodrigues, F. Carvalho

    1989-01-01

    The automated magazines are beeing used at industry more and more. One of the problems related with the automation of a Store House is the identification of the products envolved. Already used for stock management, the Bar Codes allows an easy way to identify one product. Applied to automated magazines, the bar codes allows a great variety of items in a small code. In order to be used by the national producers of automated magazines, a devoted laser scanner has been develloped. The Prototype uses an He-Ne laser whose beam scans a field angle of 75 degrees at 16 Hz. The scene reflectivity is transduced by a photodiode into an electrical signal, which is then binarized. This digital signal is the input of the decodifying program. The machine is able to see barcodes and to decode the information. A parallel interface allows the comunication with the central unit, which is responsible for the management of automated magazine.

  10. 21 CFR 892.1220 - Fluorescent scanner.

    Code of Federal Regulations, 2010 CFR

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Fluorescent scanner. 892.1220 Section 892.1220 Food and Drugs FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES (CONTINUED) MEDICAL DEVICES RADIOLOGY DEVICES Diagnostic Devices § 892.1220 Fluorescent scanner. (a) Identification....