Science.gov

Sample records for addition calibration method

  1. Inspection system calibration methods

    DOEpatents

    Deason, Vance A.; Telschow, Kenneth L.

    2004-12-28

    An inspection system calibration method includes producing two sideband signals of a first wavefront; interfering the two sideband signals in a photorefractive material, producing an output signal therefrom having a frequency and a magnitude; and producing a phase modulated operational signal having a frequency different from the output signal frequency, a magnitude, and a phase modulation amplitude. The method includes determining a ratio of the operational signal magnitude to the output signal magnitude, determining a ratio of a 1st order Bessel function of the operational signal phase modulation amplitude to a 0th order Bessel function of the operational signal phase modulation amplitude, and comparing the magnitude ratio to the Bessel function ratio.

  2. Dilution standard addition calibration: A practical calibration strategy for multiresidue organic compounds determination.

    PubMed

    Martins, Manoel L; Rizzetti, Tiele M; Kemmerich, Magali; Saibt, Nathália; Prestes, Osmar D; Adaime, Martha B; Zanella, Renato

    2016-08-19

    Among calibration approaches for organic compounds determination in complex matrices, external calibration, based in solutions of the analytes in solvent or in blank matrix extracts, is the most applied approach. Although matrix matched calibration (MMC) can compensates the matrix effects, it does not compensate low recovery results. In this way, standard addition (SA) and procedural standard calibration (PSC) are usual alternatives, despite they generate more sample and/or matrix blanks consumption need, extra sample preparations and higher time and costs. Thus, the goal of this work was to establish a fast and efficient calibration approach, the diluted standard addition calibration (DSAC), based on successive dilutions of a spiked blank sample. In order to evaluate the proposed approach, solvent calibration (SC), MMC, PSC and DSAC were applied to evaluate recovery results of grape blank samples spiked with 66 pesticides. Samples were extracted with the acetate QuEChERS method and the compounds determined by ultra-high performance liquid chromatography tandem mass spectrometry (UHPLC-MS/MS). Results indicated that low recovery results for some pesticides were compensated by both PSC and DSAC approaches. Considering recoveries from 70 to 120% with RSD <20% as adequate, DSAC presented 83%, 98% and 100% of compounds meeting this criteria for the spiking levels 10, 50 and 100μgkg(-1), respectively. PSC presented same results (83%, 98% and 100%), better than those obtained by MMC (79%, 95% and 97%) and by SC (62%, 70% and 79%). The DSAC strategy showed to be suitable for calibration of multiresidue determination methods, producing adequate results in terms of trueness and is easier and faster to perform than other approaches.

  3. Method for calibrating mass spectrometers

    DOEpatents

    Anderson, Gordon A [Benton City, WA; Brands, Michael D [Richland, WA; Bruce, James E [Schwenksville, PA; Pasa-Tolic, Ljiljana [Richland, WA; Smith, Richard D [Richland, WA

    2002-12-24

    A method whereby a mass spectra generated by a mass spectrometer is calibrated by shifting the parameters used by the spectrometer to assign masses to the spectra in a manner which reconciles the signal of ions within the spectra having equal mass but differing charge states, or by reconciling ions having known differences in mass to relative values consistent with those known differences. In this manner, the mass spectrometer is calibrated without the need for standards while allowing the generation of a highly accurate mass spectra by the instrument.

  4. FTIR Calibration Methods and Issues

    NASA Astrophysics Data System (ADS)

    Perron, Gaetan

    Over the past 10 years, several space-borne FTIR missions were launched for atmospheric research, environmental monitoring and meteorology. One can think of the Michelson Interferometer for Passive Atmospheric Sounding (MIPAS) launched by the European Space Agency, the Atmospheric Chemistry Experiment (ACE) launched by the Canadian Space Agency, the Tropospheric Emission Spectrometer (TES) launched by NASA and the Infrared Atmospheric Sounding Interferometer (IASI) launched by Eumetsat in Europe. Others are near to be launched, namely the Cross-track Infrared Sounder (CrIS) from the Integrated Program Of- fice in the United States and the Thermal And Near infrared Sensor for carbon Observation (TANSO) from the Japan Aerospace Exploration Agency. Moreover, several missions under definition foresee the use of this technology as sensor, e.g. Meteosat Third Generation (MTG), Eumetsat Polar System (EPS) and the Premier mission, one of the six candidates of the next ESA Earth Explorer Core Mission. In order to produce good quality products, calibration is essential. Calibrated data is the output of three main sub-systems that are tightly coupled: the instrument, the calibration targets and the level 1B processor. Calibration requirements must be carefully defined and propagated to each sub-system. Often, they are carried out by different parties which add to the complexity. Under budget and schedule pressure, some aspects are sometimes neglected and jeopardized final quality. For space-borne FTIR, level 1B outputs are spectra that are radiometrically, spectrally calibrated and geolocated. Radiometric calibration means to assign an intensity value in units to the y-axis. Spectral calibration means to assign to the x-axis the proper frequency value in units. Finally, geolocated means to assign a target position over the earth geoid i.e. longitude, latitude and altitude. This paper will present calibration methods and issues related to space-borne FTIR missions, e.g. two

  5. MODIS Radiometric Calibration Program, Methods and Results

    NASA Technical Reports Server (NTRS)

    Xiong, Xiaoxiong; Guenther, Bruce; Angal, Amit; Barnes, William; Salomonson, Vincent; Sun, Junqiang; Wenny, Brian

    2012-01-01

    As a key instrument for NASA s Earth Observing System (EOS), the Moderate Resolution Imaging Spectroradiometer (MODIS) has made significant contributions to the remote sensing community with its unprecedented amount of data products continuously generated from its observations and freely distributed to users worldwide. MODIS observations, covering spectral regions from visible (VIS) to long-wave infrared (LWIR), have enabled a broad range of research activities and applications for studies of the earth s interactive system of land, oceans, and atmosphere. In addition to extensive pre-launch measurements, developed to characterize sensor performance, MODIS carries a set of on-board calibrators (OBC) that can be used to track on-orbit changes of various sensor characteristics. Most importantly, dedicated and continuous calibration efforts have been made to maintain sensor data quality. This paper provides an overview of the MODIS calibration program, on-orbit calibration activities, methods, and performance. Key calibration results and lessons learned from the MODIS calibration effort are also presented in this paper.

  6. A Comparison of Two Balance Calibration Model Building Methods

    NASA Technical Reports Server (NTRS)

    DeLoach, Richard; Ulbrich, Norbert

    2007-01-01

    Simulated strain-gage balance calibration data is used to compare the accuracy of two balance calibration model building methods for different noise environments and calibration experiment designs. The first building method obtains a math model for the analysis of balance calibration data after applying a candidate math model search algorithm to the calibration data set. The second building method uses stepwise regression analysis in order to construct a model for the analysis. Four balance calibration data sets were simulated in order to compare the accuracy of the two math model building methods. The simulated data sets were prepared using the traditional One Factor At a Time (OFAT) technique and the Modern Design of Experiments (MDOE) approach. Random and systematic errors were introduced in the simulated calibration data sets in order to study their influence on the math model building methods. Residuals of the fitted calibration responses and other statistical metrics were compared in order to evaluate the calibration models developed with different combinations of noise environment, experiment design, and model building method. Overall, predicted math models and residuals of both math model building methods show very good agreement. Significant differences in model quality were attributable to noise environment, experiment design, and their interaction. Generally, the addition of systematic error significantly degraded the quality of calibration models developed from OFAT data by either method, but MDOE experiment designs were more robust with respect to the introduction of a systematic component of the unexplained variance.

  7. Langley method of calibrating UV filter radiometers

    NASA Astrophysics Data System (ADS)

    Slusser, James; Gibson, James; Bigelow, David; Kolinski, Donald; Disterhoft, Patrick; Lantz, Kathleen; Beaubien, Arthur

    2000-02-01

    The Langley method of calibrating UV multifilter shadow band radiometers (UV-MFRSR) is explored in this paper. This method has several advantages over the traditional standard lamp calibrations: the Sun is a free, universally available, and very constant source, and nearly continual automated field calibrations can be made. Although 20 or so Langley events are required for an accurate calibration, the radiometer remains in the field during calibration. Difficulties arise as a result of changing ozone optical depth during the Langley event and the breakdown of the Beer-Lambert law over the finite filter band pass since optical depth changes rapidly with wavelength. The Langley calibration of the radiometers depends critically upon the spectral characterization of each channel and on the wavelength and absolute calibration of the extraterrestrial spectrum used. Results of Langley calibrations for two UV-MFRSRs at Mauna Loa, Hawaii were compared to calibrations using two National Institute of Standards and Technology (NIST) traceable lamps. The objectives of this study were to compare Langley calibration factors with those from standard lamps and to compare field-of-view effects. The two radiometers were run simultaneously: one on a Sun tracker and the other in the conventional shadow-band configuration. Both radiometers were calibrated with two secondary 1000 W lamp, and later, the spectral response functions of the channels were measured. The ratio of Langley to lamp calibration factors for the seven channels from 300 nm to 368 nm using the shadow-band configuration ranged from 0.988 to 1.070. The estimated uncertainty in accuracy of the Langley calibrations ranged from ±3.8% at 300 nm to ±2.1% at 368 nm. For all channels calibrated with Central Ultraviolet Calibration Facility (CUCF) lamps the estimated uncertainty was ±2.5% for all channels.

  8. Study on self-calibration angle encoder using simulation method

    NASA Astrophysics Data System (ADS)

    Wang, Yan; Xue, Zi; Huang, Yao; Wang, Xiaona

    2016-01-01

    The angle measurement technology is very important in precision manufacture, optical industry, aerospace, aviation and navigation, etc. Further, the angle encoder, which uses concept `subdivision of full circle (2π rad=360°)' and transforms the angle into number of electronic pulse, is the most common instrument for angle measurement. To improve the accuracy of the angle encoder, a novel self-calibration method was proposed that enables the angle encoder to calibrate itself without angle reference. An angle deviation curve among 0° to 360° was simulated with equal weights Fourier components for the study of the self-calibration method. In addition, a self-calibration algorithm was used in the process of this deviation curve. The simulation result shows the relationship between the arrangement of multi-reading heads and the Fourier components distribution of angle encoder deviation curve. Besides, an actual self-calibration angle encoder was calibrated by polygon angle standard in national institute of metrology, China. The experiment result indicates the actual self-calibration effect on the Fourier components distribution of angle encoder deviation curve. In the end, the comparison, which is between the simulation self-calibration result and the experiment self-calibration result, reflects good consistency and proves the reliability of the self-calibration angle encoder.

  9. Automatic alignment method for calibration of hydrometers

    NASA Astrophysics Data System (ADS)

    Lee, Y. J.; Chang, K. H.; Chon, J. C.; Oh, C. Y.

    2004-04-01

    This paper presents a new method to automatically align specific scale-marks for the calibration of hydrometers. A hydrometer calibration system adopting the new method consists of a vision system, a stepping motor, and software to control the system. The vision system is composed of a CCD camera and a frame grabber, and is used to acquire images. The stepping motor moves the camera, which is attached to the vessel containing a reference liquid, along the hydrometer. The operating program has two main functions: to process images from the camera to find the position of the horizontal plane and to control the stepping motor for the alignment of the horizontal plane with a particular scale-mark. Any system adopting this automatic alignment method is a convenient and precise means of calibrating a hydrometer. The performance of the proposed method is illustrated by comparing the calibration results using the automatic alignment method with those obtained using the manual method.

  10. Simple transfer calibration method for a Cimel Sun-Moon photometer: calculating lunar calibration coefficients from Sun calibration constants.

    PubMed

    Li, Zhengqiang; Li, Kaitao; Li, Donghui; Yang, Jiuchun; Xu, Hua; Goloub, Philippe; Victori, Stephane

    2016-09-20

    The Cimel new technologies allow both daytime and nighttime aerosol optical depth (AOD) measurements. Although the daytime AOD calibration protocols are well established, accurate and simple nighttime calibration is still a challenging task. Standard lunar-Langley and intercomparison calibration methods both require specific conditions in terms of atmospheric stability and site condition. Additionally, the lunar irradiance model also has some known limits on its uncertainty. This paper presents a simple calibration method that transfers the direct-Sun calibration constant, V0,Sun, to the lunar irradiance calibration coefficient, CMoon. Our approach is a pure calculation method, independent of site limits, e.g., Moon phase. The method is also not affected by the lunar irradiance model limitations, which is the largest error source of traditional calibration methods. Besides, this new transfer calibration approach is easy to use in the field since CMoon can be obtained directly once V0,Sun is known. Error analysis suggests that the average uncertainty of CMoon over the 440-1640 nm bands obtained with the transfer method is 2.4%-2.8%, depending on the V0,Sun approach (Langley or intercomparison), which is comparable with that of lunar-Langley approach, theoretically. In this paper, the Sun-Moon transfer and the Langley methods are compared based on site measurements in Beijing, and the day-night measurement continuity and performance are analyzed.

  11. New method for spectrofluorometer monochromator wavelength calibration.

    PubMed

    Paladini, A A; Erijman, L

    1988-09-01

    A method is presented for wavelength calibration of spectrofluorometer monochromators. It is based on the distortion that the characteristic absorption bands of glass filters (holmium or didymium oxide), commonly used for calibration of spectrophotometers, introduce in the emitted fluorescence of fluorophores like indole, diphenyl hexatriene, xylene or rhodamine 6G. Those filters or a well characterized absorber with sharp bands like benzene vapor can be used for the same purpose. The wavelength calibration accuracy obtained with this method is better than 0.1 nm, and requires no modification in the geometry of the spectrofluorometer sample compartment.

  12. A Method to Test Model Calibration Techniques

    SciTech Connect

    Judkoff, Ron; Polly, Ben; Neymark, Joel

    2016-08-26

    This paper describes a method for testing model calibration techniques. Calibration is commonly used in conjunction with energy retrofit audit models. An audit is conducted to gather information about the building needed to assemble an input file for a building energy modeling tool. A calibration technique is used to reconcile model predictions with utility data, and then the 'calibrated model' is used to predict energy savings from a variety of retrofit measures and combinations thereof. Current standards and guidelines such as BPI-2400 and ASHRAE-14 set criteria for 'goodness of fit' and assume that if the criteria are met, then the calibration technique is acceptable. While it is logical to use the actual performance data of the building to tune the model, it is not certain that a good fit will result in a model that better predicts post-retrofit energy savings. Therefore, the basic idea here is that the simulation program (intended for use with the calibration technique) is used to generate surrogate utility bill data and retrofit energy savings data against which the calibration technique can be tested. This provides three figures of merit for testing a calibration technique, 1) accuracy of the post-retrofit energy savings prediction, 2) closure on the 'true' input parameter values, and 3) goodness of fit to the utility bill data. The paper will also discuss the pros and cons of using this synthetic surrogate data approach versus trying to use real data sets of actual buildings.

  13. EMI- data calibration using different electrical methods

    NASA Astrophysics Data System (ADS)

    Zschornack, L.; Nüsch, A.; Werban, U.; Dietrich, P.

    2011-12-01

    EMI-measurements were used for a wide range of soil investigations and research including water content monitoring, the derivation and classification of different soil types or measurements of salinity. Thus, the EMI-method became a common and widely accepted tool for soil scientists for many years. In addition, several companies provide different devices for a wide range of field applications and all types of investigation. However, the method suffers from a major disadvantage regarding the comparability between measurements obtained under different conditions (e.g., different measurement devices or varying weather conditions). The measured data values cannot be stated as absolute values which is a critical prerequisite for an accurate and proper application of EMI especially at larger scales (e.g., landscape scale). In addition, measurement drifts either induced from changing weather conditions or device dependent errors can provide conductivity variations which do not reflect the natural and real distribution and variability of (soil) conductivity. The use of reference lines and repeated measurements holds the potential to detect and eliminate previously unrecognized drifts in the data. That implies that different measurements (spatial and temporal separated from each other) will be more comparable but still do not provide absolute values. In this regard the best solution that allows for the derivation of absolute conductivity values is to calibrate the EMI-data using known conductivities gathered with other electrical methods. For this purpose a series of test measurements was performed to investigate which electrical method is most feasible for a calibration of EMI-data. The measurements were conducted at the floodplain of the river Mulde in Saxony (Germany). The field site is characterized by a heterogeneous sequence of river deposits including buried back water channels which causes significant variations in conductivities. We choose a reference profile with

  14. Design and calibration of zero-additional-phase SPIDER

    SciTech Connect

    Baum, Peter; Riedle, Eberhard

    2005-09-01

    Zero-additional-phase spectral phase interferometry for direct electric field reconstruction (ZAP-SPIDER) is a novel technique for measuring the temporal shape and phase of ultrashort optical pulses directly at the interaction point of a spectroscopic experiment. The scheme is suitable for an extremely wide wavelength region from the ultraviolet to the near infrared. We present a comprehensive description of the experimental setup and design guidelines to effectively apply the technique to various wavelengths and pulse durations. The calibration of the setup and procedures to check the consistency of the measurement are discussed in detail. We show experimental data for various center wavelengths and pulse durations down to 7 fs to verify the applicability to a wide range of pulse parameters.

  15. Method of Calibrating a Force Balance

    NASA Technical Reports Server (NTRS)

    Parker, Peter A. (Inventor); Rhew, Ray D. (Inventor); Johnson, Thomas H. (Inventor); Landman, Drew (Inventor)

    2015-01-01

    A calibration system and method utilizes acceleration of a mass to generate a force on the mass. An expected value of the force is calculated based on the magnitude and acceleration of the mass. A fixture is utilized to mount the mass to a force balance, and the force balance is calibrated to provide a reading consistent with the expected force determined for a given acceleration. The acceleration can be varied to provide different expected forces, and the force balance can be calibrated for different applied forces. The acceleration may result from linear acceleration of the mass or rotational movement of the mass.

  16. A simple method to calibrate intensities of photographic slit spectrograms

    NASA Astrophysics Data System (ADS)

    Vogt, N.; Barrera, L. H.

    1985-07-01

    A wavelength-dependent intensity calibration of photographic spectrograms can be obtained through the spectrograph without any additional equipment beyond a simple neutral density filter of known transparency. This filter is introduced in the focal plane of the telescope covering part of the spectrograph slit. Exposure of the comparison lamps through the entire slit yields a calibration plate which shows a well defined density jump within each line. From the height of this jump (for many lines of widely ranging strengths) the characteristic curve can be derived. The method is described and compared to the classical calibration method with a tube sensitometer.

  17. Method of biodosimeter calibration for orbital flight

    NASA Astrophysics Data System (ADS)

    Vladislav, Petrov

    A biodosimetry method, based on estimation of an absorbed dose on the basis of assessment of unstable aberration frequency in the lymphocytes of human's peripheral blood is used broadly in ground conditions for analysis of accidental exposure to personal and citizens. A calibration curve giving the relationship between aberration frequency (generally dicentrics and centric rings) and an absorbed dose in blood samples is used for assessment of crewmember exposure. As a rule gamma - rays corresponding to character of exposure in such accidents are used for these goals. At the same time the space radiation fields are formed mainly by charged particles for which the character of effect on the body cells and tissues differs strongly from that of gamma - rays. As biodosimetry is a relative method of dose measurement it is necessary to obtain a calibration curve corresponding to the conditions in which the measurements will be performed. That is a calibration curve for space application should give a relationship between aberration frequency and a dose formed by radiation field equal to that on the spacecraft trajectory. The report contains a method of obtaining a calibration curve for a case of an orbital flight on the ISS trajectory. The radiobiological basis of the method consists of relationships between chromosomal aberration frequency in human blood lymphocytes and an absorbed dose of protons with four energies (50 MeV, 150 MeV, 400 MeV, 625 MeV) obtained in the accelerator's experiments. Due to the fact that we had experimental data only for protons the calibration curve was obtained for the proton component on the ISS orbit which is mainly formed by trapped protons. Dose spectrum for this energy distribution of protons was calculated and weighting coefficients for taking into account the input of dose of protons with various energies in forming total frequency of chromosomal aberrations were obtained on its basis. The procedure of obtaining such weighting

  18. New Method of Calibrating IRT Models.

    ERIC Educational Resources Information Center

    Jiang, Hai; Tang, K. Linda

    This discussion of new methods for calibrating item response theory (IRT) models looks into new optimization procedures, such as the Genetic Algorithm (GA) to improve on the use of the Newton-Raphson procedure. The advantages of using a global optimization procedure like GA is that this kind of procedure is not easily affected by local optima and…

  19. Automatic calibration method for plenoptic camera

    NASA Astrophysics Data System (ADS)

    Luan, Yinsen; He, Xing; Xu, Bing; Yang, Ping; Tang, Guomao

    2016-04-01

    An automatic calibration method is proposed for a microlens-based plenoptic camera. First, all microlens images on the white image are searched and recognized automatically based on digital morphology. Then, the center points of microlens images are rearranged according to their relative position relationships. Consequently, the microlens images are located, i.e., the plenoptic camera is calibrated without the prior knowledge of camera parameters. Furthermore, this method is appropriate for all types of microlens-based plenoptic cameras, even the multifocus plenoptic camera, the plenoptic camera with arbitrarily arranged microlenses, or the plenoptic camera with different sizes of microlenses. Finally, we verify our method by the raw data of Lytro. The experiments show that our method has higher intelligence than the methods published before.

  20. Improved calibration method for depolarization lidar measurement.

    PubMed

    Liu, Bo; Wang, Zhien

    2013-06-17

    An improved calibration method for lidar depolarization measurement is described. With this method the system constants including the electronic gain ratio of the parallel and perpendicular channels, the optical reflectance and transmission parameters of the polarizing beam splitter, and the linear polarization ratio of the emitting laser beam can be determined conveniently by using lidar measurements with a half-wave plate oriented at selected angles.

  1. Calibration method helps in seismic velocity interpretation

    SciTech Connect

    Guzman, C.E.; Davenport, H.A.; Wilhelm, R.

    1997-11-03

    Acoustic velocities derived from seismic reflection data, when properly calibrated to subsurface measurements, help interpreters make pure velocity predictions. A method of calibrating seismic to measured velocities has improved interpretation of subsurface features in the Gulf of Mexico. In this method, the interpreter in essence creates a kind of gauge. Properly calibrated, the gauge enables the interpreter to match predicted velocities to velocities measured at wells. Slow-velocity zones are of special interest because they sometimes appear near hydrocarbon accumulations. Changes in velocity vary in strength with location; the structural picture is hidden unless the variations are accounted for by mapping in depth instead of time. Preliminary observations suggest that the presence of hydrocarbons alters the lithology in the neighborhood of the trap; this hydrocarbon effect may be reflected in the rock velocity. The effect indicates a direct use of seismic velocity in exploration. This article uses the terms seismic velocity and seismic stacking velocity interchangeably. It uses ground velocity, checkshot average velocity, and well velocity interchangeably. Interval velocities are derived from seismic stacking velocities or well average velocities; they refer to velocities of subsurface intervals or zones. Interval travel time (ITT) is the reciprocal of interval velocity in microseconds per foot.

  2. Research on calibration method of relative infrared radiometer

    NASA Astrophysics Data System (ADS)

    Yang, Sen; Li, Chengwei

    2016-02-01

    The Relative Infrared Radiometer (RIR) is commonly used to measure the irradiance of the Infrared Target Simulator (ITS), and the calibration of the RIR is central for the measurement accuracy. RIR calibration is conventionally performed using the Radiance Based (RB) calibration method or Irradiance Based (IB) calibration method, and the relationship between the radiation of standard source and the response of RIR is determined by curve fitting. One limitation existing in the calibration of RIR is the undesirable calibration voltage fluctuation in single measurement or in the reproducibility measurement, which reduces the calibration reproducibility and irradiance measurement accuracy. To address this limitation, the Equivalent Blackbody Temperature Based (EBTB) calibration method is proposed for the calibration of RIR. The purpose of this study is to compare the proposed EBTB calibration method with conventional RB and IB calibration methods. The comparison and experiment results have shown that the EBTB calibration method is not only able to provide comparable correlation between radiation and response to other calibration methods (IB and RB) in the irradiance measurement but also reduces the influence of calibration voltage fluctuation on the irradiance measurement result, which improves the calibration reproducibility and irradiance measurement accuracy.

  3. A simple calibration method for mechanically braked cycle ergometers.

    PubMed

    Van Praagh, E; Bedu, M; Roddier, P; Coudert, J

    1992-01-01

    The calibration of cycle ergometers should be checked regularly. Some studies have shown calibration errors of more than 40%. A simple, inexpensive calibrating method for mechanically braked cycle ergometers was developed and tried out on a new type of ergocycle. The cycle ergometer was elevated and the crank replaced by a pulley fitted to the shaft. The crank speed (rpm) increased linearly as a function of time when different masses were applied on the pulley. For a given braking force on the cycle ergometer, different accelerations corresponding to the increased pulley forces could be measured. When extrapolating for zero acceleration, it was possible to determine a "limit-force" which allowed the system to be in equilibrium. Additional force creates motion. The same experiments were repeated with increasing braking forces. Using the differently sized gear sprockets of the transmission system, it was possible to calculate the actual force, including all the resistances. The actual force found by the calibrating method was then compared with the indicated force proposed by the manufacturer. With increasing forces, the relative errors decreased from 9.6 to 2.9%. The cycle ergometer calibrated by this technique meets the standards recommended in exercise physiology.

  4. A Novel Camera Calibration Method Based on Polar Coordinate

    PubMed Central

    Gai, Shaoyan; Da, Feipeng; Fang, Xu

    2016-01-01

    A novel calibration method based on polar coordinate is proposed. The world coordinates are expressed in the form of polar coordinates, which are converted to world coordinates in the calibration process. In the beginning, the calibration points are obtained in polar coordinates. By transformation between polar coordinates and rectangular coordinates, the points turn into form of rectangular coordinates. Then, the points are matched with the corresponding image coordinates. At last, the parameters are obtained by objective function optimization. By the proposed method, the relationships between objects and cameras are expressed in polar coordinates easily. It is suitable for multi-camera calibration. Cameras can be calibrated with fewer points. The calibration images can be positioned according to the location of cameras. The experiment results demonstrate that the proposed method is an efficient calibration method. By the method, cameras are calibrated conveniently with high accuracy. PMID:27798651

  5. A digital calibration method for synthetic aperture radar systems

    NASA Technical Reports Server (NTRS)

    Larson, Richard W.; Jackson, P. L.; Kasischke, Eric S.

    1988-01-01

    A basic method to calibrate imagery from synthetic aperture radar (SAR) systems is presented. SAR images are calibrated by monitoring all the terms of the radar equation. This procedure includes the use of both external (calibrated reference reflectors) and internal (system-generated calibration signals) sources to monitor the total SAR system transfer function. To illustrate the implementation of the procedure, two calibrated SAR images (X-band, 3.2-cm wavelength) are presented, along with the radar cross-section measurements of specific scenes within each image. The sources of error within the SAR image calibration procedure are identified.

  6. A dynamic calibration method for the pressure transducer

    NASA Astrophysics Data System (ADS)

    Wang, Zhongyu; Wang, Zhuoran; Li, Qiang

    2016-01-01

    Pressure transducer is widely used in the field of industry. A calibrated pressure transducer can increase the performance of precision instruments in the closed mechanical relationship. Calibration is the key to ensure the pressure transducer with a high precision and dynamic characteristic. Unfortunately, the current calibration method can usually be used in the laboratory with a good condition and only one pressure transducer can be calibrated at each time. Therefore the calibration efficiency is hard to meet the requirement of modern industry with high efficiency. A dynamic and fast calibration technology with a calibration device and a corresponding data processing method is proposed in this paper. Firstly, the pressure transducer to be calibrated is placed in the small cavity chamber. The calibration process only contains a single loop. The outputs of each calibrated transducer are recorded automatically by the control terminal. Secondly, LabView programming is used for the information acquisition and data processing. The performance of the repeatability and nonlinear indicators can be figured out directly. At last the pressure transducers are calibrated simultaneously in the experiment to verify the suggested calibration technology. The experimental result shows this method can be used to calibrate the pressure transducer in the practical engineering measurement.

  7. Optimal Multicomponent Analysis Using the Generalized Standard Addition Method.

    ERIC Educational Resources Information Center

    Raymond, Margaret; And Others

    1983-01-01

    Describes an experiment on the simultaneous determination of chromium and magnesium by spectophotometry modified to include the Generalized Standard Addition Method computer program, a multivariate calibration method that provides optimal multicomponent analysis in the presence of interference and matrix effects. Provides instructions for…

  8. Method for calibration of plutonium NDA

    SciTech Connect

    Lemming, J.F.; Campbell, A.R.; Rodenburg, W.W.

    1980-01-01

    Calibration materials characterized by calorimetric assay can be a practical alternative to synthetic standards for the calibration of plutonium nondestructive assay. Calorimetric assay is an effective measurement system for the characterization because: it can give an absolute assay from first principles when the isotopic composition is known, it is insensitive to most matrix effects, and its traceability to international measurement systems has been demonstrated.

  9. Pulsed Electric Propulsion Thrust Stand Calibration Method

    NASA Technical Reports Server (NTRS)

    Wong, Andrea R.; Polzin, Kurt A.; Pearson, J. Boise

    2011-01-01

    The evaluation of the performance of any propulsion device requires the accurate measurement of thrust. While chemical rocket thrust is typically measured using a load cell, the low thrust levels associated with electric propulsion (EP) systems necessitate the use of much more sensitive measurement techniques. The design and development of electric propulsion thrust stands that employ a conventional hanging pendulum arm connected to a balance mechanism consisting of a secondary arm and variable linkage have been reported in recent publications by Polzin et al. These works focused on performing steady-state thrust measurements and employed a static analysis of the thrust stand response. In the present work, we present a calibration method and data that will permit pulsed thrust measurements using the Variable Amplitude Hanging Pendulum with Extended Range (VAHPER) thrust stand. Pulsed thrust measurements are challenging in general because the pulsed thrust (impulse bit) occurs over a short timescale (typically 1 micros to 1 millisecond) and cannot be resolved directly. Consequently, the imparted impulse bit must be inferred through observation of the change in thrust stand motion effected by the pulse. Pulsed thrust measurements have typically only consisted of single-shot operation. In the present work, we discuss repetition-rate pulsed thruster operation and describe a method to perform these measurements. The thrust stand response can be modeled as a spring-mass-damper system with a repetitive delta forcing function to represent the impulsive action of the thruster.

  10. Calibration method for a central catadioptric-perspective camera system.

    PubMed

    He, Bingwei; Chen, Zhipeng; Li, Youfu

    2012-11-01

    A central catadioptric-perspective camera system is widely used nowadays. A critical problem is that current calibration methods cannot determine the extrinsic parameters between the central catadioptric camera and a perspective camera effectively. We present a novel calibration method for a central catadioptric-perspective camera system, in which the central catadioptric camera has a hyperbolic mirror. Two cameras are used to capture images of one calibration pattern at different spatial positions. A virtual camera is constructed at the origin of the central catadioptric camera and faced toward the calibration pattern. The transformation between the virtual camera and the calibration pattern could be computed first and the extrinsic parameters between the central catadioptric camera and the calibration pattern could be obtained. Three-dimensional reconstruction results of the calibration pattern show a high accuracy and validate the feasibility of our method.

  11. Calibration methods for rotating shadowband irradiometers and evaluation of calibration duration

    NASA Astrophysics Data System (ADS)

    Jessen, W.; Wilbert, S.; Nouri, B.; Geuder, N.; Fritz, H.

    2015-10-01

    Resource assessment for Concentrated Solar Power (CSP) needs accurate Direct Normal Irradiance (DNI) measurements. An option for such measurement campaigns are Rotating Shadowband Irradiometers (RSIs) with a thorough calibration. Calibration of RSIs and Si-sensors in general is complex because of the inhomogeneous spectral response of such sensors and incorporates the use of several correction functions. A calibration for a given atmospheric condition and air mass might not work well for a different condition. This paper covers procedures and requirements for two calibration methods for the calibration of Rotating Shadowband Irradiometers. The necessary duration of acquisition of test measurements is examined in regard to the site specific conditions at Plataforma Solar de Almeria (PSA) in Spain. Data sets of several long-term calibration periods from PSA are used to evaluate the deviation of results from calibrations with varying duration from the long-term result. The findings show that seasonal changes of environmental conditions are causing small but noticeable fluctuation of calibration results. Certain periods (i.e. November to January and April to May) show a higher likelihood of particularly adverse calibration results. These effects can partially be compensated by increasing the inclusions of measurements from outside these periods. Consequently, the duration of calibrations at PSA can now be selected depending on the time of the year in which measurements are commenced.

  12. Unexplored Indoors method for pyranometers calibration traceable to SI

    NASA Astrophysics Data System (ADS)

    Castillo-Matadamas, H. A.; Molina-Vazquez, J. C.; Quintero-Torres, R.

    2015-01-01

    A method to calibrate pyranometers with direct traceability to the International System of Units (SI) is presented, the method use an electrically calibrated pyroelectric detector (ECPR) as standard and offers numerous advantages over outdoors conventional calibration methods, such as reducing the uncertainty from the reference standard and the final uncertainty of the sensitivity coefficient of the calibrated pyranometer; the measurement uncertainty achieved with this method at normal irradiance is 2.1% for a coverage factor k = 2 and could be reduce if one reduces the uncertainty level of the reference standard.

  13. Calibration methods for rotating shadowband irradiometers and optimizing the calibration duration

    NASA Astrophysics Data System (ADS)

    Jessen, Wilko; Wilbert, Stefan; Nouri, Bijan; Geuder, Norbert; Fritz, Holger

    2016-04-01

    Resource assessment for concentrated solar power (CSP) needs accurate direct normal irradiance (DNI) measurements. An option for such measurement campaigns is the use of thoroughly calibrated rotating shadowband irradiometers (RSIs). Calibration of RSIs and Si-sensors is complex because of the inhomogeneous spectral response of these sensors and incorporates the use of several correction functions. One calibration for a given atmospheric condition and air mass might not be suitable under different conditions. This paper covers procedures and requirements of two calibration methods for the calibration of rotating shadowband irradiometers. The necessary duration of acquisition of test measurements is examined with regard to the site-specific conditions at Plataforma Solar de Almería (PSA) in Spain. Seven data sets of long-term test measurements were collected. For each data set, calibration results of varying durations were compared to its respective long-term result. Our findings show that seasonal changes of environmental conditions are causing small but noticeable fluctuation of calibration results. Calibration results within certain periods (i.e. November to January and April to May) show a higher likelihood of deviation. These effects can partially be attenuated by including more measurements from outside these periods. Consequently, the duration of calibrations at PSA can now be selected depending on the time of year in which measurements commence.

  14. Calibration Method of an Ultrasonic System for Temperature Measurement

    PubMed Central

    Zhou, Chao; Wang, Yueke; Qiao, Chunjie; Dai, Weihua

    2016-01-01

    System calibration is fundamental to the overall accuracy of the ultrasonic temperature measurement, and it is basically involved in accurately measuring the path length and the system latency of the ultrasonic system. This paper proposes a method of high accuracy system calibration. By estimating the time delay between the transmitted signal and the received signal at several different temperatures, the calibration equations are constructed, and the calibrated results are determined with the use of the least squares algorithm. The formulas are deduced for calculating the calibration uncertainties, and the possible influential factors are analyzed. The experimental results in distilled water show that the calibrated path length and system latency can achieve uncertainties of 0.058 mm and 0.038 μs, respectively, and the temperature accuracy is significantly improved by using the calibrated results. The temperature error remains within ±0.04°C consistently, and the percentage error is less than 0.15%. PMID:27788252

  15. Theoretical foundation, methods, and criteria for calibrating human vibration models using frequency response functions

    PubMed Central

    Dong, Ren G.; Welcome, Daniel E.; McDowell, Thomas W.; Wu, John Z.

    2015-01-01

    While simulations of the measured biodynamic responses of the whole human body or body segments to vibration are conventionally interpreted as summaries of biodynamic measurements, and the resulting models are considered quantitative, this study looked at these simulations from a different angle: model calibration. The specific aims of this study are to review and clarify the theoretical basis for model calibration, to help formulate the criteria for calibration validation, and to help appropriately select and apply calibration methods. In addition to established vibration theory, a novel theorem of mechanical vibration is also used to enhance the understanding of the mathematical and physical principles of the calibration. Based on this enhanced understanding, a set of criteria was proposed and used to systematically examine the calibration methods. Besides theoretical analyses, a numerical testing method is also used in the examination. This study identified the basic requirements for each calibration method to obtain a unique calibration solution. This study also confirmed that the solution becomes more robust if more than sufficient calibration references are provided. Practically, however, as more references are used, more inconsistencies can arise among the measured data for representing the biodynamic properties. To help account for the relative reliabilities of the references, a baseline weighting scheme is proposed. The analyses suggest that the best choice of calibration method depends on the modeling purpose, the model structure, and the availability and reliability of representative reference data. PMID:26740726

  16. Integrated microfluidic device for serum biomarker quantitation using either standard addition or a calibration curve.

    PubMed

    Yang, Weichun; Sun, Xiuhua; Wang, Hsiang-Yu; Woolley, Adam T

    2009-10-01

    Detection and accurate quantitation of biomarkers such as alpha-fetoprotein (AFP) can be a key aspect of early stage cancer diagnosis. Microfluidic devices provide attractive analysis capabilities, including low sample and reagent consumption, as well as short assay times. However, to date microfluidic analyzers have relied almost exclusively on calibration curves for sample quantitation, which can be problematic for complex mixtures such as human serum. We have fabricated integrated polymer microfluidic systems that can quantitatively determine fluorescently labeled AFP in human serum using either the method of standard addition or a calibration curve. Our microdevices couple an immunoaffinity purification step with rapid microchip electrophoresis separation in a laser-induced fluorescence detection system, all under automated voltage control in a miniaturized polymer microchip. In conjunction with laser-induced fluorescence detection, these systems can quantify AFP at approximately 1 ng/mL levels in approximately 10 microL of human serum in a few tens of minutes. Our polymer microdevices have been applied in determining AFP in spiked serum samples. These integrated microsystems offer excellent potential for rapid, simple, and accurate biomarker quantitation in a point-of-care setting.

  17. 40 CFR 80.157 - Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2013 CFR

    2013-07-01

    ... ADDITIVES Detergent Gasoline § 80.157 Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements. This section contains requirements for automated detergent blending facilities and hand-blending detergent facilities. All gasolines and all PRC intended for use in gasoline must...

  18. 40 CFR 80.157 - Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... ADDITIVES Detergent Gasoline § 80.157 Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements. This section contains requirements for automated detergent blending facilities and hand-blending detergent facilities. All gasolines and all PRC intended for use in gasoline must...

  19. 40 CFR 80.170 - Volumetric additive reconciliation (VAR), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... ADDITIVES Detergent Gasoline § 80.170 Volumetric additive reconciliation (VAR), equipment calibration, and recordkeeping requirements. This section contains requirements for automated detergent blending facilities and hand-blending detergent facilities. All gasoline and all PRC intended for use in gasoline must...

  20. 40 CFR 80.170 - Volumetric additive reconciliation (VAR), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2013 CFR

    2013-07-01

    ... ADDITIVES Detergent Gasoline § 80.170 Volumetric additive reconciliation (VAR), equipment calibration, and recordkeeping requirements. This section contains requirements for automated detergent blending facilities and hand-blending detergent facilities. All gasoline and all PRC intended for use in gasoline must...

  1. Simultaneous multi-headed imager geometry calibration method

    DOEpatents

    Tran, Vi-Hoa; Meikle, Steven Richard; Smith, Mark Frederick

    2008-02-19

    A method for calibrating multi-headed high sensitivity and high spatial resolution dynamic imaging systems, especially those useful in the acquisition of tomographic images of small animals. The method of the present invention comprises: simultaneously calibrating two or more detectors to the same coordinate system; and functionally correcting for unwanted detector movement due to gantry flexing.

  2. Uncertainty Analysis of Inertial Model Attitude Sensor Calibration and Application with a Recommended New Calibration Method

    NASA Technical Reports Server (NTRS)

    Tripp, John S.; Tcheng, Ping

    1999-01-01

    Statistical tools, previously developed for nonlinear least-squares estimation of multivariate sensor calibration parameters and the associated calibration uncertainty analysis, have been applied to single- and multiple-axis inertial model attitude sensors used in wind tunnel testing to measure angle of attack and roll angle. The analysis provides confidence and prediction intervals of calibrated sensor measurement uncertainty as functions of applied input pitch and roll angles. A comparative performance study of various experimental designs for inertial sensor calibration is presented along with corroborating experimental data. The importance of replicated calibrations over extended time periods has been emphasized; replication provides independent estimates of calibration precision and bias uncertainties, statistical tests for calibration or modeling bias uncertainty, and statistical tests for sensor parameter drift over time. A set of recommendations for a new standardized model attitude sensor calibration method and usage procedures is included. The statistical information provided by these procedures is necessary for the uncertainty analysis of aerospace test results now required by users of industrial wind tunnel test facilities.

  3. Effect of calibration method on Tekscan sensor accuracy.

    PubMed

    Brimacombe, Jill M; Wilson, David R; Hodgson, Antony J; Ho, Karen C T; Anglin, Carolyn

    2009-03-01

    Tekscan pressure sensors are used in biomechanics research to measure joint contact loads. While the overall accuracy of these sensors has been reported previously, the effects of different calibration algorithms on sensor accuracy have not been compared. The objectives of this validation study were to determine the most appropriate calibration method supplied in the Tekscan program software and to compare its accuracy to the accuracy obtained with two user-defined calibration protocols. We evaluated the calibration accuracies for test loads within the low range, high range, and full range of the sensor. Our experimental setup used materials representing those found in standard prosthetic joints, i.e., metal against plastic. The Tekscan power calibration was the most accurate of the algorithms provided with the system software, with an overall rms error of 2.7% of the tested sensor range, whereas the linear calibrations resulted in an overall rms error of up to 24% of the tested range. The user-defined ten-point cubic calibration was almost five times more accurate, on average, than the power calibration over the full range, with an overall rms error of 0.6% of the tested range. The user-defined three-point quadratic calibration was almost twice as accurate as the Tekscan power calibration, but was sensitive to the calibration loads used. We recommend that investigators design their own calibration curves not only to improve accuracy but also to understand the range(s) of highest error and to choose the optimal points within the expected sensing range for calibration. Since output and sensor nonlinearity depend on the experimental protocol (sensor type, interface shape and materials, sensor range in use, loading method, etc.), sensor behavior should be investigated for each different application.

  4. Progress Report of CNES Activities Regarding the Absolute Calibration Method

    DTIC Science & Technology

    2010-11-01

    several receivers (Ashtech Z12-T, Septentrio PolaRx2, and Dicom GTR50) and a GNSS signal simulator (Spirent 4760) according to the temperature and...laboratories, Ashtech Z12- T, Septentrio PolaRx2, and Dicom GTR50, can be calibrated with the absolute method [6,8]. The last works concerned the...Ashtech, Septentrio, and Dicom receiver calibrations. Table 2. Uncertainty of the different receiver calibrations. Uncertainty Source

  5. Calibration method for video and radiation imagers

    DOEpatents

    Cunningham, Mark F.; Fabris, Lorenzo; Gee, Timothy F.; Goddard, Jr., James S.; Karnowski, Thomas P.; Ziock, Klaus-peter

    2011-07-05

    The relationship between the high energy radiation imager pixel (HERIP) coordinate and real-world x-coordinate is determined by a least square fit between the HERIP x-coordinate and the measured real-world x-coordinates of calibration markers that emit high energy radiation imager and reflect visible light. Upon calibration, a high energy radiation imager pixel position may be determined based on a real-world coordinate of a moving vehicle. Further, a scale parameter for said high energy radiation imager may be determined based on the real-world coordinate. The scale parameter depends on the y-coordinate of the moving vehicle as provided by a visible light camera. The high energy radiation imager may be employed to detect radiation from moving vehicles in multiple lanes, which correspondingly have different distances to the high energy radiation imager.

  6. Study on the calibration method of metrological performance of ring laser gyroscope

    NASA Astrophysics Data System (ADS)

    Qiao, Dan; Xue, Zi; Huang, Yao

    2016-10-01

    Ring Laser Gyroscope (RLG) is a kind of typical inertial device widely used in navigation field. Owing its recent advances, RLG is proposed to be used as angle measuring instrument with highly dynamic performance and loose installation requirement. To prove the reliability of RLG's angular measuring result, the metrological performance of RLG was analyzed through calibration. A new calibration method was studied to separate RLG's additional error resource such as bias drift. The special calibrating procedure was designed and the corresponding calibration system was set up. The calibration result shows that the RLG angular measuring system has measuring deviation within +/-0.4″ and repeatability within +/-0.2″. The study proposes an effective calibration method of RLG's metrological performance and proves the possibility of RLG's application in high accuracy in-situ metrology field.

  7. System and method for calibrating a rotary absolute position sensor

    NASA Technical Reports Server (NTRS)

    Davis, Donald R. (Inventor); Permenter, Frank Noble (Inventor); Radford, Nicolaus A (Inventor)

    2012-01-01

    A system includes a rotary device, a rotary absolute position (RAP) sensor generating encoded pairs of voltage signals describing positional data of the rotary device, a host machine, and an algorithm. The algorithm calculates calibration parameters usable to determine an absolute position of the rotary device using the encoded pairs, and is adapted for linearly-mapping an ellipse defined by the encoded pairs to thereby calculate the calibration parameters. A method of calibrating the RAP sensor includes measuring the rotary position as encoded pairs of voltage signals, linearly-mapping an ellipse defined by the encoded pairs to thereby calculate the calibration parameters, and calculating an absolute position of the rotary device using the calibration parameters. The calibration parameters include a positive definite matrix (A) and a center point (q) of the ellipse. The voltage signals may include an encoded sine and cosine of a rotary angle of the rotary device.

  8. Method and apparatus for calibrating a particle emissions monitor

    DOEpatents

    Flower, William L.; Renzi, Ronald F.

    1998-07-07

    The instant invention discloses method and apparatus for calibrating particulate emissions monitors, in particular, and sampling probes, in general, without removing the instrument from the system being monitored. A source of one or more specific metals in aerosol (either solid or liquid) or vapor form is housed in the instrument. The calibration operation is initiated by moving a focusing lens, used to focus a light beam onto an analysis location and collect the output light response, from an operating position to a calibration position such that the focal point of the focusing lens is now within a calibration stream issuing from a calibration source. The output light response from the calibration stream can be compared to that derived from an analysis location in the operating position to more accurately monitor emissions within the emissions flow stream.

  9. Method and apparatus for calibrating a particle emissions monitor

    DOEpatents

    Flower, W.L.; Renzi, R.F.

    1998-07-07

    The invention discloses a method and apparatus for calibrating particulate emissions monitors, in particular, sampling probes, and in general, without removing the instrument from the system being monitored. A source of one or more specific metals in aerosol (either solid or liquid) or vapor form is housed in the instrument. The calibration operation is initiated by moving a focusing lens, used to focus a light beam onto an analysis location and collect the output light response, from an operating position to a calibration position such that the focal point of the focusing lens is now within a calibration stream issuing from a calibration source. The output light response from the calibration stream can be compared to that derived from an analysis location in the operating position to more accurately monitor emissions within the emissions flow stream. 6 figs.

  10. A Bionic Polarization Navigation Sensor and Its Calibration Method

    PubMed Central

    Zhao, Huijie; Xu, Wujian

    2016-01-01

    The polarization patterns of skylight which arise due to the scattering of sunlight in the atmosphere can be used by many insects for deriving compass information. Inspired by insects’ polarized light compass, scientists have developed a new kind of navigation method. One of the key techniques in this method is the polarimetric sensor which is used to acquire direction information from skylight. In this paper, a polarization navigation sensor is proposed which imitates the working principles of the polarization vision systems of insects. We introduce the optical design and mathematical model of the sensor. In addition, a calibration method based on variable substitution and non-linear curve fitting is proposed. The results obtained from the outdoor experiments provide support for the feasibility and precision of the sensor. The sensor’s signal processing can be well described using our mathematical model. A relatively high degree of accuracy in polarization measurement can be obtained without any error compensation. PMID:27527171

  11. Method of Noncontact Calibration of the Robotic Ultrasonic Tomograph

    NASA Astrophysics Data System (ADS)

    Borikov, V. N.; Galtseva, O. V.; Filippov, G. A.

    2016-01-01

    The method of calibration of robotic ultrasonic tomograph with the construction of the trajectory of movement of the robot-manipulator on the object of control by using 3D- scanner is described. This method can significantly accelerate the process of calibration of tomograph and prevent possible displacement of the object during calibration. The algorithm of transition from use of a contact method of calibration of the tomograph to noncontact calibration is offered. Experimental data of application of this algorithm show a positive result: the time of research of object considerably decreases. Results of researches prove the practical relevance of the presented work and high efficiency of application of robotic ultrasonic tomography for nondestructive testing of objects of different forms.

  12. Novel crystal timing calibration method based on total variation

    NASA Astrophysics Data System (ADS)

    Yu, Xingjian; Isobe, Takashi; Watanabe, Mitsuo; Liu, Huafeng

    2016-11-01

    A novel crystal timing calibration method based on total variation (TV), abbreviated as ‘TV merge’, has been developed for a high-resolution positron emission tomography (PET) system. The proposed method was developed for a system with a large number of crystals, it can provide timing calibration at the crystal level. In the proposed method, the timing calibration process was formulated as a linear problem. To robustly optimize the timing resolution, a TV constraint was added to the linear equation. Moreover, to solve the computer memory problem associated with the calculation of the timing calibration factors for systems with a large number of crystals, the merge component was used for obtaining the crystal level timing calibration values. Compared with other conventional methods, the data measured from a standard cylindrical phantom filled with a radioisotope solution was sufficient for performing a high-precision crystal-level timing calibration. In this paper, both simulation and experimental studies were performed to demonstrate the effectiveness and robustness of the TV merge method. We compare the timing resolutions of a 22Na point source, which was located in the field of view (FOV) of the brain PET system, with various calibration techniques. After implementing the TV merge method, the timing resolution improved from 3.34 ns at full width at half maximum (FWHM) to 2.31 ns FWHM.

  13. Scatterometer-Calibrated Stability Verification Method

    NASA Technical Reports Server (NTRS)

    McWatters, Dalia A.; Cheetham, Craig M.; Huang, Shouhua; Fischman, Mark A.; CHu, Anhua J.; Freedman, Adam P.

    2011-01-01

    The requirement for scatterometer-combined transmit-receive gain variation knowledge is typically addressed by sampling a portion of the transmit signal, attenuating it with a known-stable attenuation, and coupling it into the receiver chain. This way, the gain variations of the transmit and receive chains are represented by this loop-back calibration signal, and can be subtracted from the received remote radar echo. Certain challenges are presented by this process, such as transmit and receive components that are outside of this loop-back path and are not included in this calibration, as well as the impracticality for measuring the transmit and receive chains stability and post fabrication separately, without the resulting measurement errors from the test set up exceeding the requirement for the flight instrument. To cover the RF stability design challenge, the portions of the scatterometer that are not calibrated by the loop-back, (e.g., attenuators, switches, diplexers, couplers, and coaxial cables) are tightly thermally controlled, and have been characterized over temperature to contribute less than 0.05 dB of calibration error over worst-case thermal variation. To address the verification challenge, including the components that are not calibrated by the loop-back, a stable fiber optic delay line (FODL) was used to delay the transmitted pulse, and to route it into the receiver. In this way, the internal loopback signal amplitude variations can be compared to the full transmit/receive external path, while the flight hardware is in the worst-case thermal environment. The practical delay for implementing the FODL is 100 s. The scatterometer pulse width is 1 ms so a test mode was incorporated early in the design phase to scale the 1 ms pulse at 100-Hz pulse repetition interval (PRI), by a factor of 18, to be a 55 s pulse with 556 s PRI. This scaling maintains the duty cycle, thus maintaining a representative thermal state for the RF components. The FODL consists

  14. Method for large-range structured light system calibration.

    PubMed

    An, Yatong; Bell, Tyler; Li, Beiwen; Xu, Jing; Zhang, Song

    2016-11-20

    Structured light system calibration often requires the usage of a calibration target with a similar size as the field of view (FOV), which brings challenges to a large-range structured light system calibration since fabricating large calibration targets is difficult and expensive. This paper presents a large-range system calibration method that does not need a large calibration target. The proposed method includes two stages: (1) accurately calibrate intrinsics (i.e., focal lengths and principle points) at a near range where both the camera and projector are out of focus, and (2) calibrate the extrinsic parameters (translation and rotation) from camera to projector with the assistance of a low-accuracy, large-range three-dimensional (3D) sensor (e.g., Microsoft Kinect). We have developed a large-scale 3D shape measurement system with a FOV of 1120  mm×1900  mm×1000  mm. Experiments demonstrate our system can achieve measurement accuracy as high as 0.07 mm with a standard deviation of 0.80 mm by measuring a 304.8 mm diameter sphere. As a comparison, Kinect V2 only achieved mean error of 0.80 mm with a standard deviation of 3.41 mm for the FOV of measurement.

  15. Verification of the ISO calibration method for field pyranometers under tropical sky conditions

    NASA Astrophysics Data System (ADS)

    Janjai, Serm; Tohsing, Korntip; Pattarapanitchai, Somjet; Detkhon, Pasakorn

    2017-02-01

    Field pyranomters need to be annually calibrated and the International Organization for Standardization (ISO) has defined a standard method (ISO 9847) for calibrating these pyranometers. According to this standard method for outdoor calibration, the field pyranometers have to be compared to a reference pyranometer for the period of 2 to 14 days, depending on sky conditions. In this work, the ISO 9847 standard method was verified under tropical sky conditions. To verify the standard method, calibration of field pyranometers was conducted at a tropical site located in Nakhon Pathom (13.82o N, 100.04o E), Thailand under various sky conditions. The conditions of the sky were monitored by using a sky camera. The calibration results for different time periods used for the calibration under various sky conditions were analyzed. It was found that the calibration periods given by this standard method could be reduced without significant change in the final calibration result. In addition, recommendation and discussion on the use of this standard method in the tropics were also presented.

  16. Noncontact method for calibration of lateral forces in scanning force microscopy.

    PubMed

    Wagner, Kyle; Cheng, Peng; Vezenov, Dmitri

    2011-04-19

    This paper describes a noncontact calibration procedure for lateral force microscopy in air and liquids. The procedure is based on the observation that the sensitivity of a force microscope may be calibrated using the raw thermal noise spectrum of the cantilever and its known spring constant, which can be found from the same uncalibrated thermal noise spectrum using Sader's method (Rev. Sci. Instrum.1999, 70, 3967-3969). In addition to the power spectrum of the cantilever thermal noise, this noncontact calibration method only requires knowledge of the plan view dimensions of the cantilever that could be measured using an optical microscope. This method is suitable for in situ force calibration even in viscous fluids through a two-step calibration procedure, where the cantilever thermal spectra are captured both in air and in the desired liquid. The lateral calibration performed with the thermal noise technique agrees well with sensitivity values obtained by the wedge calibration procedure. The approach examined in this paper allows for complete calibration of normal and lateral forces without contacting the surface, eliminating the possibility for any tip damage or contamination during calibration.

  17. Developing new online calibration methods for multidimensional computerized adaptive testing.

    PubMed

    Chen, Ping; Wang, Chun; Xin, Tao; Chang, Hua-Hua

    2017-02-01

    Multidimensional computerized adaptive testing (MCAT) has received increasing attention over the past few years in educational measurement. Like all other formats of CAT, item replenishment is an essential part of MCAT for its item bank maintenance and management, which governs retiring overexposed or obsolete items over time and replacing them with new ones. Moreover, calibration precision of the new items will directly affect the estimation accuracy of examinees' ability vectors. In unidimensional CAT (UCAT) and cognitive diagnostic CAT, online calibration techniques have been developed to effectively calibrate new items. However, there has been very little discussion of online calibration in MCAT in the literature. Thus, this paper proposes new online calibration methods for MCAT based upon some popular methods used in UCAT. Three representative methods, Method A, the 'one EM cycle' method and the 'multiple EM cycles' method, are generalized to MCAT. Three simulation studies were conducted to compare the three new methods by manipulating three factors (test length, item bank design, and level of correlation between coordinate dimensions). The results showed that all the new methods were able to recover the item parameters accurately, and the adaptive online calibration designs showed some improvements compared to the random design under most conditions.

  18. A stochastic approximation method for assigning values to calibrators.

    PubMed

    Schlain, B

    1998-04-01

    A new procedure is provided for transferring analyte concentration values from a reference material to production calibrators. This method is robust to calibration curve-fitting errors and can be accomplished using only one instrument and one set of reagents. An easily implemented stochastic approximation algorithm iteratively finds the appropriate analyte level of a standard prepared from a reference material that will yield the same average signal response as the new production calibrator. Alternatively, a production bulk calibrator material can be iteratively adjusted to give the same average signal response as some prespecified, fixed reference standard. In either case, the outputted value assignment of the production calibrator is the analyte concentration of the reference standard in the final iteration of the algorithm. Sample sizes are statistically determined as functions of known within-run signal response precisions and user-specified accuracy tolerances.

  19. Evaluation of two gas-dilution methods for instrument calibration

    NASA Technical Reports Server (NTRS)

    Evans, A., Jr.

    1977-01-01

    Two gas dilution methods were evaluated for use in the calibration of analytical instruments used in air pollution studies. A dual isotope fluorescence carbon monoxide analyzer was used as the transfer standard. The methods are not new but some modifications are described. The rotary injection gas dilution method was found to be more accurate than the closed loop method. Results by the two methods differed by 5 percent. This could not be accounted for by the random errors in the measurements. The methods avoid the problems associated with pressurized cylinders. Both methods have merit and have found a place in instrument calibration work.

  20. Method calibration of the model 13145 infrared target projectors

    NASA Astrophysics Data System (ADS)

    Huang, Jianxia; Gao, Yuan; Han, Ying

    2014-11-01

    The SBIR Model 13145 Infrared Target Projectors ( The following abbreviation Evaluation Unit ) used for characterizing the performances of infrared imaging system. Test items: SiTF, MTF, NETD, MRTD, MDTD, NPS. Infrared target projectors includes two area blackbodies, a 12 position target wheel, all reflective collimator. It provide high spatial frequency differential targets, Precision differential targets imaged by infrared imaging system. And by photoelectricity convert on simulate signal or digital signal. Applications software (IR Windows TM 2001) evaluate characterizing the performances of infrared imaging system. With regards to as a whole calibration, first differently calibration for distributed component , According to calibration specification for area blackbody to calibration area blackbody, by means of to amend error factor to calibration of all reflective collimator, radiance calibration of an infrared target projectors using the SR5000 spectral radiometer, and to analyze systematic error. With regards to as parameter of infrared imaging system, need to integrate evaluation method. According to regulation with -GJB2340-1995 General specification for military thermal imaging sets -testing parameters of infrared imaging system, the results compare with results from Optical Calibration Testing Laboratory . As a goal to real calibration performances of the Evaluation Unit.

  1. A Method to Test Model Calibration Techniques: Preprint

    SciTech Connect

    Judkoff, Ron; Polly, Ben; Neymark, Joel

    2016-09-01

    This paper describes a method for testing model calibration techniques. Calibration is commonly used in conjunction with energy retrofit audit models. An audit is conducted to gather information about the building needed to assemble an input file for a building energy modeling tool. A calibration technique is used to reconcile model predictions with utility data, and then the 'calibrated model' is used to predict energy savings from a variety of retrofit measures and combinations thereof. Current standards and guidelines such as BPI-2400 and ASHRAE-14 set criteria for 'goodness of fit' and assume that if the criteria are met, then the calibration technique is acceptable. While it is logical to use the actual performance data of the building to tune the model, it is not certain that a good fit will result in a model that better predicts post-retrofit energy savings. Therefore, the basic idea here is that the simulation program (intended for use with the calibration technique) is used to generate surrogate utility bill data and retrofit energy savings data against which the calibration technique can be tested. This provides three figures of merit for testing a calibration technique, 1) accuracy of the post-retrofit energy savings prediction, 2) closure on the 'true' input parameter values, and 3) goodness of fit to the utility bill data. The paper will also discuss the pros and cons of using this synthetic surrogate data approach versus trying to use real data sets of actual buildings.

  2. Two laboratory methods for the calibration of GPS speed meters

    NASA Astrophysics Data System (ADS)

    Bai, Yin; Sun, Qiao; Du, Lei; Yu, Mei; Bai, Jie

    2015-01-01

    The set-ups of two calibration systems are presented to investigate calibration methods of GPS speed meters. The GPS speed meter calibrated is a special type of high accuracy speed meter for vehicles which uses Doppler demodulation of GPS signals to calculate the measured speed of a moving target. Three experiments are performed: including simulated calibration, field-test signal replay calibration, and in-field test comparison with an optical speed meter. The experiments are conducted at specific speeds in the range of 40-180 km h-1 with the same GPS speed meter as the device under calibration. The evaluation of measurement results validates both methods for calibrating GPS speed meters. The relative deviations between the measurement results of the GPS-based high accuracy speed meter and those of the optical speed meter are analyzed, and the equivalent uncertainty of the comparison is evaluated. The comparison results justify the utilization of GPS speed meters as reference equipment if no fewer than seven satellites are available. This study contributes to the widespread use of GPS-based high accuracy speed meters as legal reference equipment in traffic speed metrology.

  3. Method to calibrate fission chambers in Campbelling mode

    SciTech Connect

    Benoit Geslot; Troy C. Unruh; Philippe Filliatre; Christian Jammes; Jacques Di Salvo; Stéphane Bréaud; Jean-François Villard

    2011-06-01

    Fission chambers are neutron detectors which are widely used to instrument experimental reactors such as material testing reactors or zero power reactors. In the presence of a high level mixed gamma and neutron flux, fission chambers can be operated in Campbelling mode (also known as 'fluctuation mode' or 'mean square voltage mode') to provide reliable and precise neutron related measurements. Fission chamber calibration in Campbelling mode (in terms of neutron flux) is usually done empirically using a calibrated reference detector. A major drawback of this method is that calibration measurements have to be performed in a neutron environment very similar to the one in which the calibrated detector will be used afterwards. What we propose here is a different approach based on characterizing the fission chamber response in terms of fission rate. This way, the detector calibration coefficient is independent from the neutron spectrum and can be determined prior to the experiment. The fissile deposit response to the neutron spectrum can then be assessed independently by other means (experimental or numerical). In this paper, the response of CEA made miniature fission chambers in Campbelling mode is studied. We use a theoretical model of the signal to calculate the calibration coefficient. Input parameters of the model come from statistical distribution of individual pulses. Supporting measurements have been made in the CEA Cadarache zero power reactor MINERVE. Results are compared to an empirical Campbelling mode calibration.

  4. Method of calibrating an interferometer and reducing its systematic noise

    NASA Technical Reports Server (NTRS)

    Hammer, Philip D. (Inventor)

    1997-01-01

    Methods of operation and data analysis for an interferometer so as to eliminate the errors contributed by non-responsive or unstable pixels, interpixel gain variations that drift over time, and spurious noise that would otherwise degrade the operation of the interferometer are disclosed. The methods provide for either online or post-processing calibration. The methods apply prescribed reversible transformations that exploit the physical properties of interferograms obtained from said interferometer to derive a calibration reference signal for subsequent treatment of said interferograms for interpixel gain variations. A self-consistent approach for treating bad pixels is incorporated into the methods.

  5. Site characterization for calibration of radiometric sensors using vicarious method

    NASA Astrophysics Data System (ADS)

    Parihar, Shailesh; Rathore, L. S.; Mohapatra, M.; Sharma, A. K.; Mitra, A. K.; Bhatla, R.; Singh, R. S.; Desai, Yogdeep; Srivastava, Shailendra S.

    2016-05-01

    Radiometric performances of earth observation satellite/sensors vary from ground pre-launch calibration campaign to post launch period extended to lifetime of the satellite due to launching vibrations. Therefore calibration is carried out worldwide through various methods throughout satellite lifetime. In India Indian Space Research Organization (ISRO) calibrates the sensor of Resourcesat-2 satellite by vicarious method. One of these vicarious calibration methods is the reflectance-based approach that is applied in this study for radiometric calibration of sensors on-board Resouresat-2 satellite. The results of ground-based measurement of atmospheric conditions and surface reflectance are made at Bap, Rajasthan Calibration/Validation (Cal/Val) site. Cal/Val observations at site were carried out with hyper-spectral Spectroradiometer covering spectral range of 350nm- 2500nm for radiometric characterization of the site. The Sunphotometer/Ozonometer for measuring the atmospheric parameters has also been used. The calibrated radiance is converted to absolute at-sensor spectral reflectance and Top-Of-Atmosphere (TOA) radiance. TOA radiance was computed using radiative transfer model `Second simulation of the satellite signal in the solar spectrum' (6S), which can accurately simulate the problems introduced by the presence of the atmosphere along the path from Sun to target (surface) to Sensor. The methodology for band averaged reflectance retrieval and spectral reflectance fitting process are described. Then the spectral reflectance and atmospheric parameters are put into 6S code to predict TOA radiance which compare with Resourcesat-2 radiance. Spectral signature and its reflectance ratio indicate the uniformity of the site. Thus the study proves that the selected site is suitable for vicarious calibration of sensor of Resourcesat-2. Further the study demonstrates the procedure for similar exercise for site selection for Cal/Val analysis of other satellite over India

  6. Method and apparatus for calibrating a linear variable differential transformer

    DOEpatents

    Pokrywka, Robert J.

    2005-01-18

    A calibration apparatus for calibrating a linear variable differential transformer (LVDT) having an armature positioned in au LVDT armature orifice, and the armature able to move along an axis of movement. The calibration apparatus includes a heating mechanism with an internal chamber, a temperature measuring mechanism for measuring the temperature of the LVDT, a fixture mechanism with an internal chamber for at least partially accepting the LVDT and for securing the LVDT within the heating mechanism internal chamber, a moving mechanism for moving the armature, a position measurement mechanism for measuring the position of the armature, and an output voltage measurement mechanism. A method for calibrating an LVDT, including the steps of: powering the LVDT; heating the LVDT to a desired temperature; measuring the position of the armature with respect to the armature orifice; and measuring the output voltage of the LVDT.

  7. An Accurate Projector Calibration Method Based on Polynomial Distortion Representation

    PubMed Central

    Liu, Miao; Sun, Changku; Huang, Shujun; Zhang, Zonghua

    2015-01-01

    In structure light measurement systems or 3D printing systems, the errors caused by optical distortion of a digital projector always affect the precision performance and cannot be ignored. Existing methods to calibrate the projection distortion rely on calibration plate and photogrammetry, so the calibration performance is largely affected by the quality of the plate and the imaging system. This paper proposes a new projector calibration approach that makes use of photodiodes to directly detect the light emitted from a digital projector. By analyzing the output sequence of the photoelectric module, the pixel coordinates can be accurately obtained by the curve fitting method. A polynomial distortion representation is employed to reduce the residuals of the traditional distortion representation model. Experimental results and performance evaluation show that the proposed calibration method is able to avoid most of the disadvantages in traditional methods and achieves a higher accuracy. This proposed method is also practically applicable to evaluate the geometric optical performance of other optical projection system. PMID:26492247

  8. Sphere-based calibration method for trinocular vision sensor

    NASA Astrophysics Data System (ADS)

    Lu, Rui; Shao, Mingwei

    2017-03-01

    A new method to calibrate a trinocular vision sensor is proposed and two main tasks are finished in this paper, i.e. to determine the transformation matrix between each two cameras and the trifocal tensor of the trinocular vision sensor. A flexible sphere target with several spherical circles is designed. As the isotropy of a sphere, trifocal tensor of the three cameras can be determined exactly from the feature on the sphere target. Then the fundamental matrix between each two cameras can be obtained. Easily, compatible rotation matrix and translation matrix can be deduced base on the singular value decomposition of the fundamental matrix. In our proposed calibration method, image points are not requested one-to-one correspondence. When image points locates in the same feature are obtained, the transformation matrix between each two cameras with the trifocal tensor of trinocular vision sensor can be determined. Experiment results show that the proposed calibration method can obtain precise results, including measurement and matching results. The root mean square error of distance is 0.026 mm with regard to the view field of about 200×200 mm and the feature matching of three images is strict. As a sphere projection is not concerned with its orientation, the calibration method is robust and with an easy operation. Moreover, our calibration method also provides a new approach to obtain the trifocal tensor.

  9. Method for out-of-focus camera calibration.

    PubMed

    Bell, Tyler; Xu, Jing; Zhang, Song

    2016-03-20

    State-of-the-art camera calibration methods assume that the camera is at least nearly in focus and thus fail if the camera is substantially defocused. This paper presents a method which enables the accurate calibration of an out-of-focus camera. Specifically, the proposed method uses a digital display (e.g., liquid crystal display monitor) to generate fringe patterns that encode feature points into the carrier phase; these feature points can be accurately recovered, even if the fringe patterns are substantially blurred (i.e., the camera is substantially defocused). Experiments demonstrated that the proposed method can accurately calibrate a camera regardless of the amount of defocusing: the focal length difference is approximately 0.2% when the camera is focused compared to when the camera is substantially defocused.

  10. Geometric calibration of a terrestrial laser scanner with local additional parameters: An automatic strategy

    NASA Astrophysics Data System (ADS)

    García-San-Miguel, D.; Lerma, J. L.

    2013-05-01

    Terrestrial laser scanning systems are steadily increasing in many fields of engineering, geoscience and architecture namely for fast data acquisition, 3-D modeling and mapping. Similarly to other precision instruments, these systems provide measurements with implicit systematic errors. Systematic errors are physically corrected by manufacturers before delivery and sporadically afterwards. The approach presented herein tackles the raw observables acquired by a laser scanner with additional parameters, a set of geometric calibration parameters that model the systematic error of the instrument to achieve the most accurate point cloud outputs, improving eventual workflow owing to less filtering, better registration and best 3D modeling. This paper presents a fully automatic strategy to calibrate geometrically terrestrial laser scanning datasets. The strategy is tested with multiple scans taken by a FARO FOCUS 3D, a phase-based terrestrial laser scanner. A calibration with local parameters for datasets is undertaken to improve the raw observables and a weighted mathematical index is proposed to select the most significant set of additional parameters. The improvements achieved are exposed, highlighting the necessity of correcting the terrestrial laser scanner before handling multiple data sets.

  11. Bond additivity corrections for quantum chemistry methods

    SciTech Connect

    C. F. Melius; M. D. Allendorf

    1999-04-01

    In the 1980's, the authors developed a bond-additivity correction procedure for quantum chemical calculations called BAC-MP4, which has proven reliable in calculating the thermochemical properties of molecular species, including radicals as well as stable closed-shell species. New Bond Additivity Correction (BAC) methods have been developed for the G2 method, BAC-G2, as well as for a hybrid DFT/MP2 method, BAC-Hybrid. These BAC methods use a new form of BAC corrections, involving atomic, molecular, and bond-wise additive terms. These terms enable one to treat positive and negative ions as well as neutrals. The BAC-G2 method reduces errors in the G2 method due to nearest-neighbor bonds. The parameters within the BAC-G2 method only depend on atom types. Thus the BAC-G2 method can be used to determine the parameters needed by BAC methods involving lower levels of theory, such as BAC-Hybrid and BAC-MP4. The BAC-Hybrid method should scale well for large molecules. The BAC-Hybrid method uses the differences between the DFT and MP2 as an indicator of the method's accuracy, while the BAC-G2 method uses its internal methods (G1 and G2MP2) to provide an indicator of its accuracy. Indications of the average error as well as worst cases are provided for each of the BAC methods.

  12. Reliably Detectable Flaw Size for NDE Methods that Use Calibration

    NASA Technical Reports Server (NTRS)

    Koshti, Ajay M.

    2017-01-01

    Probability of detection (POD) analysis is used in assessing reliably detectable flaw size in nondestructive evaluation (NDE). MIL-HDBK-1823 and associated mh1823 POD software gives most common methods of POD analysis. In this paper, POD analysis is applied to an NDE method, such as eddy current testing, where calibration is used. NDE calibration standards have known size artificial flaws such as electro-discharge machined (EDM) notches and flat bottom hole (FBH) reflectors which are used to set instrument sensitivity for detection of real flaws. Real flaws such as cracks and crack-like flaws are desired to be detected using these NDE methods. A reliably detectable crack size is required for safe life analysis of fracture critical parts. Therefore, it is important to correlate signal responses from real flaws with signal responses form artificial flaws used in calibration process to determine reliably detectable flaw size.

  13. A Comparative Study of IRT Fixed Parameter Calibration Methods

    ERIC Educational Resources Information Center

    Kim, Seonghoon

    2006-01-01

    This article provides technical descriptions of five fixed parameter calibration (FPC) methods, which were based on marginal maximum likelihood estimation via the EM algorithm, and evaluates them through simulation. The five FPC methods described are distinguished from each other by how many times they update the prior ability distribution and by…

  14. An Automated Method for Ozonesonde Calibration: New Insights

    NASA Technical Reports Server (NTRS)

    Schmidlin, F. J.; Hoegger, Bruno A.; Levrat, Gilbert; Baldwin, Tony

    2008-01-01

    An automated method for preparation of the electrochemical concentration cell (ECC) ozonesonde is presented. Development of a computer-controlled system for preparation and calibration of the ECC is an improvement over the manual preparation method, and reduces subjectivity considerably. Preparation measurements in digital form aids analysis of the ECC before release and enhances post-flight data certification. Calibration of ozonesondes over a range of ozone concentrations between 0 mPA and 30 mPA is discussed. This presentation describes the automatic system, gives examples of calibrations. The automated system enables comparison of varying potassium iodide (KI) concentrations that should allow adjustment of earlier ozonesonde data obtained with different KT concentrations used since 1970, i.e., 2, 1.5, 1, and 0.5 percent. Preliminary results indicate ECC accuracy has a strong dependence on the electrolyte concentration and should not be considered linear with altitude.

  15. Bond additivity corrections for quantum chemistry methods

    SciTech Connect

    Melius, C.F.; Allendorf, M.D.

    2000-03-23

    New bond additivity correction (BAC) methods have been developed for the G2 method, BAC-G2, as well as for a hybrid density functional theory (DFT) Moller-Plesset (MP)2 method, BAC-hybrid. These BAC methods use a new form of BAC corrections, involving atomic, molecular, and bond-wise additive terms. These terms enable one to treat positive and negative ions as well as neutrals. The BAC-G2 method reduces errors in the G2 method due to nearest-neighbor bonds. The parameters within the BAC-G2 method only depend on atom types. Thus the BAC-G2 method can be used to determine the parameters needed by BAC methods involving lower levels of theory, such as BAC-hybrid and BAC-MP4. The BAC-hybrid method is expected to scale well for large molecules. The BAC-hybrid method uses the differences between the DFT and MP2 predictions as an indication of the method's accuracy, whereas the BAC-G2 method uses its internal methods (G1 and G2MP2) to accomplish this. A statistical analysis of the error in each of the methods is presented on the basis of calculations performed for large sets (more than 120) of molecules.

  16. A stoichiometric calibration method for dual energy computed tomography

    NASA Astrophysics Data System (ADS)

    Bourque, Alexandra E.; Carrier, Jean-François; Bouchard, Hugo

    2014-04-01

    The accuracy of radiotherapy dose calculation relies crucially on patient composition data. The computed tomography (CT) calibration methods based on the stoichiometric calibration of Schneider et al (1996 Phys. Med. Biol. 41 111-24) are the most reliable to determine electron density (ED) with commercial single energy CT scanners. Along with the recent developments in dual energy CT (DECT) commercial scanners, several methods were published to determine ED and the effective atomic number (EAN) for polyenergetic beams without the need for CT calibration curves. This paper intends to show that with a rigorous definition of the EAN, the stoichiometric calibration method can be successfully adapted to DECT with significant accuracy improvements with respect to the literature without the need for spectrum measurements or empirical beam hardening corrections. Using a theoretical framework of ICRP human tissue compositions and the XCOM photon cross sections database, the revised stoichiometric calibration method yields Hounsfield unit (HU) predictions within less than ±1.3 HU of the theoretical HU calculated from XCOM data averaged over the spectra used (e.g., 80 kVp, 100 kVp, 140 kVp and 140/Sn kVp). A fit of mean excitation energy (I-value) data as a function of EAN is provided in order to determine the ion stopping power of human tissues from ED-EAN measurements. Analysis of the calibration phantom measurements with the Siemens SOMATOM Definition Flash dual source CT scanner shows that the present formalism yields mean absolute errors of (0.3 ± 0.4)% and (1.6 ± 2.0)% on ED and EAN, respectively. For ion therapy, the mean absolute errors for calibrated I-values and proton stopping powers (216 MeV) are (4.1 ± 2.7)% and (0.5 ± 0.4)%, respectively. In all clinical situations studied, the uncertainties in ion ranges in water for therapeutic energies are found to be less than 1.3 mm, 0.7 mm and 0.5 mm for protons, helium and carbon ions respectively, using a generic

  17. A Calibration Method for Wide-Field Multicolor Photometric Systems

    NASA Astrophysics Data System (ADS)

    Zhou, Xu; Chen, Jiansheng; Xu, Wen; Zhang, Mei; Jiang, Zhaoji; Zheng, Zhongyuan; Zhu, Jin

    1999-07-01

    The purpose of this paper is to present a method to self-calibrate the spectral energy distribution (SED) of objects in a survey based on the fitting of a SED library to observed multicolor photometry. We adopt, for illustrative purposes, the Vilnius and Gunn & Stryker SED libraries. The self-calibration technique can improve the quality of observations which are not taken under perfectly photometric conditions. The more passbands used for the photometry, the better the results. This technique has been applied to the BATC 15 passband CCD survey.

  18. Calibration method of absolute orientation of camera optical axis

    NASA Astrophysics Data System (ADS)

    Xu, Yong; Guo, Pengyu; Zhang, Xiaohu; Ding, Shaowen; Su, Ang; Li, Lichun

    2013-08-01

    Camera calibration is one of the most basic and important processes in optical measuring field. Generally, the objective of camera calibration is to estimate the internal and external parameters of object cameras, while the orientation error of optical axis is not included yet. Orientation error of optical axis is a important factor, which seriously affects measuring precision in high-precision measurement field, especially for those distant aerospace measurement in which object distance is much longer than focal length, that lead to magnifying the orientation errors to thousands times. In order to eliminate the influence of orientation error of camera optical axis, the imaging model of camera is analysed and established in this paper, and the calibration method is also introduced: Firstly, we analyse the reasons that cause optical axis error and its influence. Then, we find the model of optical axis orientation error and imaging model of camera basing on it's practical physical meaning. Furthermore, we derive the bundle adjustment algorithm which could compute the internal and external camera parameters and absolute orientation of camera optical axis simultaneously at high precision. In numeric simulation, we solve the camera parameters by using bundle adjustment optimization algorithm, then we correct the image points by calibration results according to the model of optical axis error, and the simulation result shows that our calibration model is reliable, effective and precise.

  19. Calibration Methods for Air Coupled Antennas - COST Action TU1208

    NASA Astrophysics Data System (ADS)

    Marecos, Vânia; Solla, Mercedes; Fontul, Simona; Pajewski, Lara

    2016-04-01

    This work focuses on the comparison of different methods for calibrating air coupled antennas: Coring, Surface Reflection Method (SRM) and Common Mid-Point (CMP) through the analysis of GPR data collected in a test site with different pavement solutions. Research activities have been carried out during a Short Term Scientific Mission (STSM) funded by the COST (European Cooperation in Science and Technology) Action TU1208 "Civil Engineering Applications of Ground Penetrating Radar" in December 2015. The use of GPR in transport infrastructures represents one of the most significant advances for obtaining continuous data along the road, with the advantage of operation at traffic speed and being a non-destructive technique. Its main application has been the evaluation of layer thickness. For the determination of layer thickness, it is necessary to know the velocity of the signal, which depends on the dielectric constant of the material, and the two-way travel time of the reflected signal that is recorded by the GPR system. The calculation of the dielectric value of the materials can be done using different approaches such as: using fixed values based on experience, laboratory determination of dielectric values, applying the SRM, performing back calculation from ground truth references such as cores and test pits, or using the CMP method. The problem with using ground truth is that it is time consuming, labour intensive and intrusive to traffic, in addition, a drill core is not necessarily representative of the whole surveyed area. Regarding the surface reflection technique, one of the problems is that it only measures the dielectric value from the layer surface and not from the whole layer. Recent works already started to address some of these challenges proposing new approaches for GPR layer thickness measurements using multiple antennas to calculate the average dielectric value of the asphalt layer, taking advantage of significant hardware improvements in GPR

  20. An MLC calibration method using a detector array

    SciTech Connect

    Simon, Thomas A.; Kahler, Darren; Simon, William E.; Fox, Christopher; Li, Jonathan; Palta, Jatinder; Liu, Chihray

    2009-10-15

    Purpose: The authors have developed a quantitative calibration method for a multileaf collimator (MLC) which measures individual leaf positions relative to the MLC backup jaw on an Elekta Synergy linear accelerator. Methods: The method utilizes a commercially available two-axis detector array (Profiler 2; Sun Nuclear Corporation, Melbourne, FL). To calibrate the MLC bank, its backup jaw is positioned at the central axis and the opposing jaw is retracted to create a half-beam configuration. The position of the backup jaws field edge is then measured with the array to obtain what is termed the radiation defined reference line. The positions of the individual leaf ends relative to this reference line are then inferred by the detector response in the leaf end penumbra. Iteratively adjusting and remeasuring the leaf end positions to within specifications completes the calibration. Using the backup jaw as a reference for the leaf end positions is based on three assumptions: (1) The leading edge of an MLC leaf bank is parallel to its backup jaw's leading edge, (2) the backup jaw position is reproducible, and (3) the measured radiation field edge created by each leaf end is representative of that leaf's position. Data from an electronic portal imaging device (EPID) were used in a similar analysis to check the results obtained with the array. Results: The relative leaf end positions measured with the array differed from those measured with the EPID by an average of 0.11 {+-}0.09 mm per leaf. The maximum leaf positional change measured with the Profiler 2 over a 3 month period was 0.51 mm. A leaf positional accuracy of {+-}0.4 mm is easily attainable through the iterative calibration process. The method requires an average of 40 min to measure both leaf banks. Conclusions: This work demonstrates that the Profiler 2 is an effective tool for efficient and quantitative MLC quality assurance and calibration.

  1. A new method to calibrate the absolute sensitivity of a soft X-ray streak camera

    NASA Astrophysics Data System (ADS)

    Yu, Jian; Liu, Shenye; Li, Jin; Yang, Zhiwen; Chen, Ming; Guo, Luting; Yao, Li; Xiao, Shali

    2016-12-01

    In this paper, we introduce a new method to calibrate the absolute sensitivity of a soft X-ray streak camera (SXRSC). The calibrations are done in the static mode by using a small laser-produced X-ray source. A calibrated X-ray CCD is used as a secondary standard detector to monitor the X-ray source intensity. In addition, two sets of holographic flat-field grating spectrometers are chosen as the spectral discrimination systems of the SXRSC and the X-ray CCD. The absolute sensitivity of the SXRSC is obtained by comparing the signal counts of the SXRSC to the output counts of the X-ray CCD. Results show that the calibrated spectrum covers the range from 200 eV to 1040 eV. The change of the absolute sensitivity in the vicinity of the K-edge of the carbon can also be clearly seen. The experimental values agree with the calculated values to within 29% error. Compared with previous calibration methods, the proposed method has several advantages: a wide spectral range, high accuracy, and simple data processing. Our calibration results can be used to make quantitative X-ray flux measurements in laser fusion research.

  2. A calibration method of the multi-channel imaging lidar

    NASA Astrophysics Data System (ADS)

    Xu, Weiming; Liu, Jun; Shu, Rong

    2014-06-01

    We design a kind of imaging LiDAR with sixteen channels, which consists of a fiber laser source, dual scanning galvanometers, range measurement circuits and information processing circuits etc. The image LiDAR provides sixteen range measurements for one laser shot and the distance accuracy of each channel is about 4cm. This paper provides a calibrate method to correct point cloud images captured with the multi-channel LiDAR. The method needs to construct different slanted planes to cover the imaging field, and establish precise plane equations in the known ground coordinates, then fit planes with point clouds data and calculate correction parameters of all channels through the error model. The image accuracy is better than 5cm processed by this calibration method.

  3. Effusion plate using additive manufacturing methods

    DOEpatents

    Johnson, Thomas Edward; Keener, Christopher Paul; Ostebee, Heath Michael; Wegerif, Daniel Gerritt

    2016-04-12

    Additive manufacturing techniques may be utilized to construct effusion plates. Such additive manufacturing techniques may include defining a configuration for an effusion plate having one or more internal cooling channels. The manufacturing techniques may further include depositing a powder into a chamber, applying an energy source to the deposited powder, and consolidating the powder into a cross-sectional shape corresponding to the defined configuration. Such methods may be implemented to construct an effusion plate having one or more channels with a curved cross-sectional geometry.

  4. New float equivalent calibration method for 2D image measuring system

    NASA Astrophysics Data System (ADS)

    Gou, Jiansong; Wang, Zhong; Lu, Ruijun; Shen, Xinlan

    2015-08-01

    Pixel equivalent is an important parameter to describe the relationship between pixels of digital images and actual size of measured piece in a 2D image measuring system. It is mainly calibrated with the standard component method, which is traditionally off-line and requires measuring conditions and attitude of devices to remain constant while measuring and calibrating. To overcome above limitations, a new calibration method is proposed in this paper which is defined as the float equivalent method. This method requires the standard component and measured piece be placed in image measuring system simultaneously. Everytime before measuring, no matter aiming at the same measuring point or not, the pixel equivalent is calibrated for this specific time, specific condition, specific measuring point, and specific object distance. This method has the advantage of reducing the influence of conditions changing on the accuracy without additional calibration equipment or operations. The steel tape verification system is taken as an example to testify the effectiveness of the method.

  5. Calibration of base flow separation methods with streamflow conductivity.

    PubMed

    Stewart, Mark; Cimino, Joseph; Ross, Mark

    2007-01-01

    The conductivity mass-balance (CMB) method can be used to calibrate analytical base flow separation methods. The principal CMB assumptions are base flow conductivity is equal to streamflow conductivity at lowest flows, runoff conductivity is equal to streamflow conductivity at highest flows, and base flow and runoff conductivities are assumed to be constants over the period of record. To test the CMB assumptions, fluid conductivities of ground water, surface runoff, and streamflow were measured during wet and dry conditions in a 12-km(2) stream basin. Ground water conductivities at wells varied an average of 6% from dry to wet conditions, while stream conductivities varied 58%. Shallow ground water conductivity varied significantly with distance from the stream, with lowest conductivities of 87 microS/cm near the divide, a maximum of 520 microS/cm 59 m from the stream, and 215 microS/cm 22 m from the stream. Runoff conductivities measured in three rain events remained nearly constant, with lower conductivities of 35 microS/cm near the divide and 50 microS/cm near the stream. The CMB method was applied to the records from 10 USGS stream-gauging stations in Texas, Kentucky, Georgia, and Florida to calibrate the USGS base flow separation technique, HYSEP, by varying the time parameter 2N*. There is a statistically significant relationship between basin areas and calibrated values of 2N*, expressed as N = 0.46A(0.44), with N in days and A in km(2). The widely accepted relationship N = 0.83A(0.2) is not valid for these basins. Other analytic methods can also be calibrated with the CMB method.

  6. A Laser Sheet Self-Calibration Method for Scanning PIV

    NASA Astrophysics Data System (ADS)

    Knutsen, Anna N.; Dawson, James R.; Lawson, John M.; Worth, Nicholas A.

    2016-11-01

    A laser sheet self-calibration method for scanning PIV has been developed to replace the current laser sheet calibration, which is complex, time consuming and very sensitive to misalignment of the optics or cameras during experiments. The new calibration method is simpler, faster and crucially more robust. The concept behind the method is to traverse a laser sheet through the measurement volume, take a series of images from two different views, and calculate the global 3D particle locations. This information is used to find the real space coordinates of the measurement volume and the orientation and width of the laser sheets. The spatial location of the particles is found by object matching and triangulation. The light intensity in the laser sheet has an approximately Gaussian shape, and the illumination of one particle which will be illuminated multiple times during the scan will thus vary as the sheet is scanned across the measurement volume. The thickness of the laser sheet is calculated by identifying the variation of illumination of the particles during a scan and fitting this to a Gaussian shaped curve, while the orientation is found using a least square fit. The accuracy of the new method will be presented with respect to both synthetic and experimental data.

  7. A linear signal transmission system calibration method of wideband GPR

    NASA Astrophysics Data System (ADS)

    Wu, Bin; Zhao, Kai; Gu, Ling-jia; Cao, Qiong; Li, Xiao-feng; Zheng, Xing-ming

    2016-09-01

    In VHF pulse Ground Penetrating Radar(GPR) system, the echo pass through the antenna and transmission line circuit, then reach the GPR receiver. Thus the reflection coefficient at the receiver sampling gate interface, which is at the end of the transmission line, is different from the real reflection coefficient of the media at the antenna interface, which could cause the GPR receiving error. The pulse GPR receiver is a wideband system that can't be simply described as traditional narrowband transmission line model. Since the GPR transmission circuit is a linear system, the linear transformation method could be used to analyze the characteristic of the GPR receiving system. A GPR receiver calibration method based on transmission line theory is proposed in this paper, which analyzes the relationship between the reflection coefficients of theory calculation at antenna interface and the measuring data by network analyzer at the sampling gate interface. Then the least square method is introduced to calibrate the transfer function of the GPR receiver transmission circuit. This calibration method can be useful in media quantitative inversion by GPR. When the reflection coefficient at the sampling gate is obtained, the real reflection coefficient of the media at the antenna interface can be easily determined.

  8. Standardization of serum cholesterol assays by use of serum calibrators and direct addition of Liebermann-Burchard reagent.

    PubMed

    Katan, M B; van der Haar, F; Kromhout, D; Schouten, F J

    1982-04-01

    Serum cholesterol concentrations of subjects in epidemiological studies were measured after direct addition of Liebermann-Burchard reagent; results were calibrated with human serum pools assayed according to Abell et al. (J. Biol. Chem. 195:357-366, 1952). Accuracy and precision were monitored for six years by analysis of internal-control pools and blind external-control pools. For various internal-control pools, the imprecision (CV) of the long-term averages of run means ranged from 0.5 to 0.9%. The within-run CV for internal control and patients' sera was about 1%. For blind control sera with different concentrations (provided by the Centers for Disease Control, Atlanta, GA, over the same period), the average difference per three-month period between the values found and the target values was usually between -0.5% and +0.7% for medium-concentration pools and between -2% and +2% for low- and high-concentration pools (extreme values: -2.4% and +2.5%). The CV per three-month period ranged from 0.6 to 2.7%. Sera from subjects on diets of high or low linoleic acid content were analyzed to study the effect of the fatty acid portion of serum cholesterol esters; the differences between values obtained with the comparison method and the direct method was insignificant on both diets. We conclude that the use of serum calibrators eliminates the bias inherent in the direct method.

  9. AVIRIS calibration using the cloud-shadow method

    NASA Technical Reports Server (NTRS)

    Carder, K. L.; Reinersman, P.; Chen, R. F.

    1993-01-01

    More than 90 percent of the signal at an ocean-viewing, satellite sensor is due to the atmosphere, so a 5 percent sensor-calibration error viewing a target that contributes but 10 percent of the signal received at the sensor may result in a target-reflectance error of more than 50 percent. Since prelaunch calibration accuracies of 5 percent are typical of space-sensor requirements, recalibration of the sensor using ground-base methods is required for low-signal target. Known target reflectance or water-leaving radiance spectra and atmospheric correction parameters are required. In this article we describe an atmospheric-correction method that uses cloud shadowed pixels in combination with pixels in a neighborhood region of similar optical properties to remove atmospheric effects from ocean scenes. These neighboring pixels can then be used as known reflectance targets for validation of the sensor calibration and atmospheric correction. The method uses the difference between water-leaving radiance values for these two regions. This allows nearly identical optical contributions to the two signals (e.g., path radiance and Fresnel-reflected skylight) to be removed, leaving mostly solar photons backscattered from beneath the sea to dominate the residual signal. Normalization by incident solar irradiance reaching the sea surface provides the remote-sensing reflectance of the ocean at the location of the neighbor region.

  10. Supplier Selection Using Weighted Utility Additive Method

    NASA Astrophysics Data System (ADS)

    Karande, Prasad; Chakraborty, Shankar

    2015-10-01

    Supplier selection is a multi-criteria decision-making (MCDM) problem which mainly involves evaluating a number of available suppliers according to a set of common criteria for choosing the best one to meet the organizational needs. For any manufacturing or service organization, selecting the right upstream suppliers is a key success factor that will significantly reduce purchasing cost, increase downstream customer satisfaction and improve competitive ability. The past researchers have attempted to solve the supplier selection problem employing different MCDM techniques which involve active participation of the decision makers in the decision-making process. This paper deals with the application of weighted utility additive (WUTA) method for solving supplier selection problems. The WUTA method, an extension of utility additive approach, is based on ordinal regression and consists of building a piece-wise linear additive decision model from a preference structure using linear programming (LP). It adopts preference disaggregation principle and addresses the decision-making activities through operational models which need implicit preferences in the form of a preorder of reference alternatives or a subset of these alternatives present in the process. The preferential preorder provided by the decision maker is used as a restriction of a LP problem, which has its own objective function, minimization of the sum of the errors associated with the ranking of each alternative. Based on a given reference ranking of alternatives, one or more additive utility functions are derived. Using these utility functions, the weighted utilities for individual criterion values are combined into an overall weighted utility for a given alternative. It is observed that WUTA method, having a sound mathematical background, can provide accurate ranking to the candidate suppliers and choose the best one to fulfill the organizational requirements. Two real time examples are illustrated to prove

  11. Calibration method of microgrid polarimeters with image interpolation.

    PubMed

    Chen, Zhenyue; Wang, Xia; Liang, Rongguang

    2015-02-10

    Microgrid polarimeters have large advantages over conventional polarimeters because of the snapshot nature and because they have no moving parts. However, they also suffer from several error sources, such as fixed pattern noise (FPN), photon response nonuniformity (PRNU), pixel cross talk, and instantaneous field-of-view (IFOV) error. A characterization method is proposed to improve the measurement accuracy in visible waveband. We first calibrate the camera with uniform illumination so that the response of the sensor is uniform over the entire field of view without IFOV error. Then a spline interpolation method is implemented to minimize IFOV error. Experimental results show the proposed method can effectively minimize the FPN and PRNU.

  12. Head-free, remote eye-gaze detection system based on pupil-corneal reflection method with easy calibration using two stereo-calibrated video cameras.

    PubMed

    Ebisawa, Yoshinobu; Fukumoto, Kiyotaka

    2013-10-01

    We have developed a pupil-corneal reflection method-based gaze detection system, which allows large head movements and achieves easy gaze calibration. This system contains two optical systems consisting of components such as a camera and a near-infrared light source attached to the camera. The light source has two concentric LED rings with different wavelengths. The inner and outer rings generate bright and dark pupil images, respectively. The pupils are detected from a difference image created by subtracting the bright and dark pupil images. The light source also generates the corneal reflection. The 3-D coordinates of the pupils are determined by the stereo matching method using two optical systems. The vector from the corneal reflection center to the pupil center in the camera image is determined as r. The angle between the line of sight and the line passing through the pupil center and the camera (light source) is denoted as θ. The relationship θ =k |r| is assumed, where k is a constant. The theory implies that head movement of the user is allowed and facilitates the gaze calibration procedure. In the automatic calibration method, k is automatically determined while the user looks around on the PC screen without fixating on any specific calibration target. In the one-point calibration method, the user is asked to fixate on one calibration target at the PC screen in order to correct the difference between the optical and visual axes. In the two-point calibration method, in order to correct the nonlinear relationship between θ and |r|, the user is asked to fixate on two targets. The experimental results show that the three proposed calibration methods improve the precision of gaze detection step by step. In addition, the average gaze error in the visual angle is less than 1° for the seven head positions of the user.

  13. Automatic Calibration Method for a Storm Water Runoff Model

    NASA Astrophysics Data System (ADS)

    Barco, J.; Wong, K. M.; Hogue, T.; Stenstrom, M. K.

    2007-12-01

    Major metropolitan areas are characterized by continuous increases in imperviousness due to urban development. Increasing imperviousness increases runoff volume and maximum rates of runoff, with generally negative consequences for natural systems. To avoid environmental degradation, new development standards often prohibit increases in total runoff volume and may limit maximum flow rates. Methods to reduce runoff volume and maximum runoff rate are required, and solutions to the problems may benefit from the use of advanced models. In this study the U.S. Storm Water Management Model (SWMM) was adapted and calibrated to the Ballona Creek watershed, a large urban catchment in Southern California. A geographic information system (GIS) was used to process the input data and generate the spatial distribution of precipitation. An optimization procedure using the Complex Method was incorporated to estimate runoff parameters, and ten storms were used for calibration and validation. The calibrated model predicted the observed outputs with reasonable accuracy. A sensitivity analysis showed the impact of the model parameters, and results were most sensitive to imperviousness and impervious depression storage and least sensitive to Manning roughness for surface flow. Optimized imperviousness was greater than imperviousness predicted from landuse information. The results demonstrate that this methodology of integrating GIS and stormwater model with a constrained optimization technique can be applied to large watersheds, and can be a useful tool to evaluate alternative strategies to reduce runoff rate and volume.

  14. New Method for Calibration for Hyperspectral Pushbroom Imaging Systems

    NASA Technical Reports Server (NTRS)

    Ryan, Robert; Olive, Dan; ONeal, Duane; Schere, Chris; Nixon, Thomas; May, Chengye; Ryan, Jim; Stanley, Tom; Witcher, Kern

    1999-01-01

    A new, easy-to-implement approach for achieving highly accurate spectral and radiometric calibration of array-based, hyperspectral pushbroom imagers is presented in this paper. The equivalence of the plane of the exit port of an integrating sphere to a Lambertian surface is utilized to provide a field-filling radiance source for the imager. Several different continuous wave lasers of various wavelengths and a quartz-tungsten-halogen lamp internally illuminate the sphere. The imager is positioned to "stare" into the port, and the resultant data cube is analyzed to determine wavelength calibrations, spectral widths of channels, radiometric characteristics, and signal-to-noise ratio, as well as an estimate of signal-to-noise performance in the field. The "smile" (geometric distortion of spectra) of the system can be quickly ascertained using this method. As the price and availability of solid state laser sources improve, this technique could gain wide acceptance.

  15. Innovative self-calibration method for accelerometer scale factor of the missile-borne RINS with fiber optic gyro.

    PubMed

    Zhang, Qian; Wang, Lei; Liu, Zengjun; Zhang, Yiming

    2016-09-19

    The calibration of an inertial measurement unit (IMU) is a key technique to improve the preciseness of the inertial navigation system (INS) for missile, especially for the calibration of accelerometer scale factor. Traditional calibration method is generally based on the high accuracy turntable, however, it leads to expensive costs and the calibration results are not suitable to the actual operating environment. In the wake of developments in multi-axis rotational INS (RINS) with optical inertial sensors, self-calibration is utilized as an effective way to calibrate IMU on missile and the calibration results are more accurate in practical application. However, the introduction of multi-axis RINS causes additional calibration errors, including non-orthogonality errors of mechanical processing and non-horizontal errors of operating environment, it means that the multi-axis gimbals could not be regarded as a high accuracy turntable. As for its application on missiles, in this paper, after analyzing the relationship between the calibration error of accelerometer scale factor and non-orthogonality and non-horizontal angles, an innovative calibration procedure using the signals of fiber optic gyro and photoelectric encoder is proposed. The laboratory and vehicle experiment results validate the theory and prove that the proposed method relaxes the orthogonality requirement of rotation axes and eliminates the strict application condition of the system.

  16. Calibration of EMI data based on different electrical methods

    NASA Astrophysics Data System (ADS)

    Nüsch, Anne-Kathrin; Werban, Ulrike; Dietrich, Peter

    2013-04-01

    The advantages of the electromagnetic induction (EMI)-method have been known to soil scientists for many years. Thus it is used for many soil investigations, ranging from salinity measurements over water content monitoring to classification of different soil types. There are several companies that provide instruments for each type of investigation. However, a major disadvantage of the method is that measurements obtained under different conditions (e.g. with different instruments, or at different times or field sites) are not easily comparable. Data values yielded when using the instruments are not absolute, which is an important prerequisite for the correct application of EMI, especially at the landscape scale. Furthermore drifts can occur, potentially caused by weather conditions or instrument errors and subsequently give results with variations in conductivities, which are not actually reflective of actual test results. With the help of reference lines and repeated measurements, drifts can be detected and eliminated. Different measurements (spatial and temporal) are more comparable, but the final corrected values are still not absolute. The best solution that allows for absolute values to be obtained is to calibrate the EMI-Data with the help of a known conductivity from other electrical methods. In a series of test measurements, we studied which electrical method is most feasible for a calibration of EMI-data. The chosen field site is situated at the floodplain of the river Mulde in Saxony (Germany). We chose a profile 100 meters in length which is very heterogeneous and crosses a buried back water channel. Results show a significant variance of conductivities. Several EMI-instruments were tested. Among these are EM38DD and EM31 devices from Geonics. These instruments are capable of investigating the subsurface to a depth of up to six meters. For the calibration process, we chose electrical resistivity tomography (ERT), Vertical Electrical Sounding (VES), and

  17. Method of Calibration for a Large Cathetometer System

    NASA Technical Reports Server (NTRS)

    Toland, Ronald

    2004-01-01

    A method of calibration has been devised for a pair of mutually orthogonal two-axis cathetometers that, when used together, yield measurements of three-dimensional positions of objects mounted on an optical bench. Each cathetometer has a horizontal travel of 1.8 m and a vertical travel of 1.2 m. The cathetometers are required to measure X, Y, and Z coordinates (see figure) to within plus or minus 0.005 in. (plus or minus 0.127 mm). Each cathetometer consists of an alignment telescope on a platform mounted on a two-dimensional translation stage. The knowledge required for calibration of each cathetometer is (1) the two-dimensional position of the cathetometer platform as a function of the electronic readouts of position encoders on the translation stage and (2) the amount of any angular misalignment (roll, pitch, and/or yaw) of the cathetometer platform as a function of the two-dimensional coordinates or the position-encoder readouts. By use of three equations derived from the applicable trigonometric relationships, the calibrated X, Y, and Z coordinates can be computed from the raw encoder readouts.

  18. A fast RCS accuracy assessment method for passive radar calibrators

    NASA Astrophysics Data System (ADS)

    Zhou, Yongsheng; Li, Chuanrong; Tang, Lingli; Ma, Lingling; Liu, QI

    2016-10-01

    In microwave radar radiometric calibration, the corner reflector acts as the standard reference target but its structure is usually deformed during the transportation and installation, or deformed by wind and gravity while permanently installed outdoor, which will decrease the RCS accuracy and therefore the radiometric calibration accuracy. A fast RCS accuracy measurement method based on 3-D measuring instrument and RCS simulation was proposed in this paper for tracking the characteristic variation of the corner reflector. In the first step, RCS simulation algorithm was selected and its simulation accuracy was assessed. In the second step, the 3-D measuring instrument was selected and its measuring accuracy was evaluated. Once the accuracy of the selected RCS simulation algorithm and 3-D measuring instrument was satisfied for the RCS accuracy assessment, the 3-D structure of the corner reflector would be obtained by the 3-D measuring instrument, and then the RCSs of the obtained 3-D structure and corresponding ideal structure would be calculated respectively based on the selected RCS simulation algorithm. The final RCS accuracy was the absolute difference of the two RCS calculation results. The advantage of the proposed method was that it could be applied outdoor easily, avoiding the correlation among the plate edge length error, plate orthogonality error, plate curvature error. The accuracy of this method is higher than the method using distortion equation. In the end of the paper, a measurement example was presented in order to show the performance of the proposed method.

  19. Various methods and developments for calibrating seismological sensors at EOST

    NASA Astrophysics Data System (ADS)

    JUND, H.; Bès de Berc, M.; Thore, J.

    2013-12-01

    Calibrating seismic sensors is crucial for knowing the quality of the sensor and generating precise dataless files. We present here three calibration methods that we have developed for the short period and broad band sensors included in the temporary and permanent seismic networks in France. First, in the case of a short-period sensor with no electronics and calibration coil, we inject a sine wave signal into the signal coil. After locking the sensor mass, we first connect a voltage generator of signal waves and a series resistor to the coil. Then, a sinusoidal signal is sent to the sensor signal coil output. Both the voltage at the terminal of the resistor, which gives an image of the intensity entering the signal coil, and the voltage at the terminal of the signal coil are measured. The frequency of the generator then varies in order to find a phase shift between both signals of π/2. The output frequency of the generator corresponds to the image of the natural frequency of the sensor. Second, in the case of all types of sensors provided with a calibration coil, we inject different signals into the calibration coil. We usually apply two signals: a step signal and a sweep (or wobble) signal. A step signal into the calibration coil is equivalent to a Dirac excitation in derived acceleration. The response to this Dirac gives the transfer function of the signal coil, derived two times and without absolute gain. We developed a field-module allowing us to always apply the same excitation to various models of seismometers, in order to compare the results from several instruments previously installed on field. A wobble signal is a signal whose frequency varies. By varying the frequency of the input signal around the sensor's natural frequency, we obtain an immediate response of the sensor in acceleration. This method is particularly suitable in order to avoid any disturbances which may modify the signal of a permanent station. Finally, for the determination of absolute

  20. [Study on the absolute spectral irradiation calibration method for far ultraviolet spectrometer in remote sensing].

    PubMed

    Yu, Lei; Lin, Guan-Yu; Chen, Bin

    2013-01-01

    The present paper studied spectral irradiation responsivities calibration method which can be applied to the far ultraviolet spectrometer for upper atmosphere remote sensing. It is difficult to realize the calibration for far ultraviolet spectrometer for many reasons. Standard instruments for far ultraviolet waveband calibration are few, the degree of the vacuum experiment system is required to be high, the stabilities of the experiment are hardly maintained, and the limitation of the far ultraviolet waveband makes traditional diffuser and the integrating sphere radiance calibration method difficult to be used. To solve these problems, a new absolute spectral irradiance calibration method was studied, which can be applied to the far ultraviolet calibration. We build a corresponding special vacuum experiment system to verify the calibration method. The light source system consists of a calibrated deuterium lamp, a vacuum ultraviolet monochromater and a collimating system. We used the calibrated detector to obtain the irradiance responsivities of it. The three instruments compose the calibration irradiance source. We used the "calibration irradiance source" to illuminate the spectrometer prototype and obtained the spectral irradiance responsivities. It realized the absolute spectral irradiance calibration for the far ultraviolet spectrometer utilizing the calibrated detector. The absolute uncertainty of the calibration is 7.7%. The method is significant for the ground irradiation calibration of the far ultraviolet spectrometer in upper atmosphere remote sensing.

  1. Selenium measurement in human plasma with Zeeman effect electrothermal atomic absorption spectrometry: sample stability and calibration method.

    PubMed

    Sabé, Rosa; Rubio, Roser; García-Beltrán, Lydia

    2003-01-01

    The dual aim of the present study is the investigation of the stability of plasma samples for selenium determination with time and temperature and the assessment of the calibration method. A comparative study is performed, using two calibration methods: standard addition to each sample and matrix matched curve. Our findings show that, in general, significant differences in the selenium content are observed when comparing the results obtained with these two calibration methods. Plasma samples stored at -20 degrees C are stable relative to the selenium content for a period of at least one year.

  2. Spinning disk calibration method and apparatus for laser Doppler velocimeter

    NASA Technical Reports Server (NTRS)

    Snyder, P. K. (Inventor)

    1986-01-01

    A method and apparatus for calibrating laser Doppler velocimeters having one or more intersecting beam pairs are described. These velocimeters measure fluid velocity by observing the light scattered by particles in the fluid stream. Moving fluid particulates are simulated by fine taut wires that are radially mounted on a disk that is rotated at a known velocity. The laser beam intersection locus is first aimed at the very center of the disk and then the disk is translated so that the locus is swept by the rotating wires. The radial distance traversed is precisely measured so that the velocity of the wires (pseudo particles) may be calculated.

  3. Method of making self-calibrated displacement measurements

    DOEpatents

    Pedersen, Herbert N.

    1977-01-01

    A method for monitoring the displacement of an object having an acoustically reflective surface at least partially submerged in an acoustically conductive medium. The reflective surface is designed to have a stepped interface responsive to an incident acoustic pulse to provide separate discrete reflected pulses to a receiving transducer. The difference in the time of flight of the reflected acoustic signals corresponds to the known step height and the time of travel of the signals to the receiving transducer provides a measure of the displacement of the object. Accordingly, the reference step length enables simultaneous calibration of each displacement measurement.

  4. Method of calibrating a fluid-level measurement system

    NASA Technical Reports Server (NTRS)

    Woodard, Stanley E. (Inventor); Taylor, Bryant D. (Inventor)

    2010-01-01

    A method of calibrating a fluid-level measurement system is provided. A first response of the system is recorded when the system's sensor(s) is (are) not in contact with a fluid of interest. A second response of the system is recorded when the system's sensor(s) is (are) fully immersed in the fluid of interest. Using the first and second responses, a plurality of expected responses of the system's sensor(s) is (are) generated for a corresponding plurality of levels of immersion of the sensor(s) in the fluid of interest.

  5. Investigation of factors affecting the heater wire method of calibrating fine wire thermocouples

    NASA Technical Reports Server (NTRS)

    Keshock, E. G.

    1972-01-01

    An analytical investigation was made of a transient method of calibrating fine wire thermocouples. The system consisted of a 10 mil diameter standard thermocouple (Pt, Pt-13% Rh) and an 0.8 mil diameter chromel-alumel thermocouple attached to a 20 mil diameter electrically heated platinum wire. The calibration procedure consisted of electrically heating the wire to approximately 2500 F within about a seven-second period in an environment approximating atmospheric conditions at 120,000 feet. Rapid periodic readout of the standard and fine wire thermocouple signals permitted a comparison of the two temperature indications. An analysis was performed which indicated that the temperature distortion at the heater wire produced by the thermocouple junctions appears to be of negligible magnitude. Consequently, the calibration technique appears to be basically sound, although several practical changes which appear desirable are presented and discussed. Additional investigation is warranted to evaluate radiation effects and transient response characteristics.

  6. Line-feature-based calibration method of structured light plane parameters for robot hand-eye system

    NASA Astrophysics Data System (ADS)

    Qi, Yuhan; Jing, Fengshui; Tan, Min

    2013-03-01

    For monocular-structured light vision measurement, it is essential to calibrate the structured light plane parameters in addition to the camera intrinsic parameters. A line-feature-based calibration method of structured light plane parameters for a robot hand-eye system is proposed. Structured light stripes are selected as calibrating primitive elements, and the robot moves from one calibrating position to another with constraint in order that two misaligned stripe lines are generated. The images of stripe lines could then be captured by the camera fixed at the robot's end link. During calibration, the equations of two stripe lines in the camera coordinate system are calculated, and then the structured light plane could be determined. As the robot's motion may affect the effectiveness of calibration, so the robot's motion constraints are analyzed. A calibration experiment and two vision measurement experiments are implemented, and the results reveal that the calibration accuracy can meet the precision requirement of robot thick plate welding. Finally, analysis and discussion are provided to illustrate that the method has a high efficiency fit for industrial in-situ calibration.

  7. Methods for Calibration of Prout-Tompkins Kinetics Parameters Using EZM Iteration and GLO

    SciTech Connect

    Wemhoff, A P; Burnham, A K; de Supinski, B; Sexton, J; Gunnels, J

    2006-11-07

    This document contains information regarding the standard procedures used to calibrate chemical kinetics parameters for the extended Prout-Tompkins model to match experimental data. Two methods for calibration are mentioned: EZM calibration and GLO calibration. EZM calibration matches kinetics parameters to three data points, while GLO calibration slightly adjusts kinetic parameters to match multiple points. Information is provided regarding the theoretical approach and application procedure for both of these calibration algorithms. It is recommended that for the calibration process, the user begin with EZM calibration to provide a good estimate, and then fine-tune the parameters using GLO. Two examples have been provided to guide the reader through a general calibrating process.

  8. Method for Ground-to-Satellite Laser Calibration System

    NASA Technical Reports Server (NTRS)

    Lukashin, Constantine (Inventor); Wielicki, Bruce A. (Inventor)

    2015-01-01

    The present invention comprises an approach for calibrating the sensitivity to polarization, optics degradation, spectral and stray light response functions of instruments on orbit. The concept is based on using an accurate ground-based laser system, Ground-to-Space Laser Calibration (GSLC), transmitting laser light to instrument on orbit during nighttime substantially clear-sky conditions. To minimize atmospheric contribution to the calibration uncertainty the calibration cycles should be performed in short time intervals, and all required measurements are designed to be relative. The calibration cycles involve ground operations with laser beam polarization and wavelength changes.

  9. Method for Ground-to-Space Laser Calibration System

    NASA Technical Reports Server (NTRS)

    Lukashin, Constantine (Inventor); Wielicki, Bruce A. (Inventor)

    2014-01-01

    The present invention comprises an approach for calibrating the sensitivity to polarization, optics degradation, spectral and stray light response functions of instruments on orbit. The concept is based on using an accurate ground-based laser system, Ground-to-Space Laser Calibration (GSLC), transmitting laser light to instrument on orbit during nighttime substantially clear-sky conditions. To minimize atmospheric contribution to the calibration uncertainty the calibration cycles should be performed in short time intervals, and all required measurements are designed to be relative. The calibration cycles involve ground operations with laser beam polarization and wavelength changes.

  10. Flight Test Results of an Angle of Attack and Angle of Sideslip Calibration Method Using Output-Error Optimization

    NASA Technical Reports Server (NTRS)

    Siu, Marie-Michele; Martos, Borja; Foster, John V.

    2013-01-01

    As part of a joint partnership between the NASA Aviation Safety Program (AvSP) and the University of Tennessee Space Institute (UTSI), research on advanced air data calibration methods has been in progress. This research was initiated to expand a novel pitot-static calibration method that was developed to allow rapid in-flight calibration for the NASA Airborne Subscale Transport Aircraft Research (AirSTAR) facility. This approach uses Global Positioning System (GPS) technology coupled with modern system identification methods that rapidly computes optimal pressure error models over a range of airspeed with defined confidence bounds. Subscale flight tests demonstrated small 2-s error bounds with significant reduction in test time compared to other methods. Recent UTSI full scale flight tests have shown airspeed calibrations with the same accuracy or better as the Federal Aviation Administration (FAA) accepted GPS 'four-leg' method in a smaller test area and in less time. The current research was motivated by the desire to extend this method for inflight calibration of angle of attack (AOA) and angle of sideslip (AOS) flow vanes. An instrumented Piper Saratoga research aircraft from the UTSI was used to collect the flight test data and evaluate flight test maneuvers. Results showed that the output-error approach produces good results for flow vane calibration. In addition, maneuvers for pitot-static and flow vane calibration can be integrated to enable simultaneous and efficient testing of each system.

  11. A SVD-based method to assess the uniqueness and accuracy of SPECT geometrical calibration.

    PubMed

    Ma, Tianyu; Yao, Rutao; Shao, Yiping; Zhou, Rong

    2009-12-01

    Geometrical calibration is critical to obtaining high resolution and artifact-free reconstructed image for SPECT and CT systems. Most published calibration methods use analytical approach to determine the uniqueness condition for a specific calibration problem, and the calibration accuracy is often evaluated through empirical studies. In this work, we present a general method to assess the characteristics of both the uniqueness and the quantitative accuracy of the calibration. The method uses a singular value decomposition (SVD) based approach to analyze the Jacobian matrix from a least-square cost function for the calibration. With this method, the uniqueness of the calibration can be identified by assessing the nonsingularity of the Jacobian matrix, and the estimation accuracy of the calibration parameters can be quantified by analyzing the SVD components. A direct application of this method is that the efficacy of a calibration configuration can be quantitatively evaluated by choosing a figure-of-merit, e.g., the minimum required number of projection samplings to achieve desired calibration accuracy. The proposed method was validated with a slit-slat SPECT system through numerical simulation studies and experimental measurements with point sources and an ultra-micro hot-rod phantom. The predicted calibration accuracy from the numerical studies was confirmed by the experimental point source calibrations at approximately 0.1 mm for both the center of rotation (COR) estimation of a rotation stage and the slit aperture position (SAP) estimation of a slit-slat collimator by an optimized system calibration protocol. The reconstructed images of a hot rod phantom showed satisfactory spatial resolution with a proper calibration and showed visible resolution degradation with artificially introduced 0.3 mm COR estimation error. The proposed method can be applied to other SPECT and CT imaging systems to analyze calibration method assessment and calibration protocol

  12. On-site calibration method for outdoor binocular stereo vision sensors

    NASA Astrophysics Data System (ADS)

    Liu, Zhen; Yin, Yang; Wu, Qun; Li, Xiaojing; Zhang, Guangjun

    2016-11-01

    Using existing calibration methods for binocular stereo vision sensors (BSVS), it is very difficult to extract target characteristic points in outdoor environments under complex light conditions. To solve the problem, an online calibration method for BSVS based a double parallel cylindrical target and a line laser projector is proposed in this paper. The intrinsic parameters of two cameras are calibrated offline. Laser strips on the double parallel cylindrical target are mediated to calibrate the configuration parameters of BSVS. The proposed method only requires images of laser strips on the target and is suitable for the calibration of BSVS in outdoor environments. The effectiveness of the proposed method is validated through physical experiments.

  13. A method of camera calibration with adaptive thresholding

    NASA Astrophysics Data System (ADS)

    Gao, Lei; Yan, Shu-hua; Wang, Guo-chao; Zhou, Chun-lei

    2009-07-01

    In order to calculate the parameters of the camera correctly, we must figure out the accurate coordinates of the certain points in the image plane. Corners are the important features in the 2D images. Generally speaking, they are the points that have high curvature and lie in the junction of different brightness regions of images. So corners detection has already widely used in many fields. In this paper we use the pinhole camera model and SUSAN corner detection algorithm to calibrate the camera. When using the SUSAN corner detection algorithm, we propose an approach to retrieve the gray difference threshold, adaptively. That makes it possible to pick up the right chessboard inner comers in all kinds of gray contrast. The experiment result based on this method was proved to be feasible.

  14. Method for in-situ calibration of electrophoretic analysis systems

    DOEpatents

    Liu, Changsheng; Zhao, Hequan

    2005-05-08

    An electrophoretic system having a plurality of separation lanes is provided with an automatic calibration feature in which each lane is separately calibrated. For each lane, the calibration coefficients map a spectrum of received channel intensities onto values reflective of the relative likelihood of each of a plurality of dyes being present. Individual peaks, reflective of the influence of a single dye, are isolated from among the various sets of detected light intensity spectra, and these can be used to both detect the number of dye components present, and also to establish exemplary vectors for the calibration coefficients which may then be clustered and further processed to arrive at a calibration matrix for the system. The system of the present invention thus permits one to use different dye sets to tag DNA nucleotides in samples which migrate in separate lanes, and also allows for in-situ calibration with new, previously unused dye sets.

  15. One step geometrical calibration method for optical coherence tomography

    NASA Astrophysics Data System (ADS)

    Díaz Díaz, Jesús; Stritzel, Jenny; Rahlves, Maik; Majdani, Omid; Reithmeier, Eduard; Ortmaier, Tobias; Roth, Bernhard

    2016-01-01

    We present a novel one-step calibration methodology for geometrical distortion correction for optical coherence tomography (OCT). A calibration standard especially designed for OCT is introduced, which consists of an array of inverse pyramidal structures. The use of multiple landmarks situated on four different height levels on the pyramids allow performing a 3D geometrical calibration. The calibration procedure itself is based on a parametric model of the OCT beam propagation. It is validated by experimental results and enables the reduction of systematic errors by more than one order of magnitude. In future, our results can improve OCT image reconstruction and interpretation for medical applications such as real time monitoring of surgery.

  16. Transmission electron microscope calibration methods for critical dimension standards

    NASA Astrophysics Data System (ADS)

    Orji, Ndubuisi G.; Dixson, Ronald G.; Garcia-Gutierrez, Domingo I.; Bunday, Benjamin D.; Bishop, Michael; Cresswell, Michael W.; Allen, Richard A.; Allgair, John A.

    2016-10-01

    One of the key challenges in critical dimension (CD) metrology is finding suitable dimensional calibration standards. The transmission electron microscope (TEM), which produces lattice-resolved images having scale traceability to the SI (International System of Units) definition of length through an atomic lattice constant, has gained wide usage in different areas of CD calibration. One such area is critical dimension atomic force microscope (CD-AFM) tip width calibration. To properly calibrate CD-AFM tip widths, errors in the calibration process must be quantified. Although the use of TEM for CD-AFM tip width calibration has been around for about a decade, there is still confusion on what should be considered in the uncertainty analysis. We characterized CD-AFM tip-width samples using high-resolution TEM and high angle annular dark field scanning TEM and two CD-AFMs that are implemented as reference measurement systems. The results are used to outline how to develop a rigorous uncertainty estimate for TEM/CD-AFM calibration, and to compare how information from the two electron microscopy modes are applied to practical CD-AFM measurements. The results also represent a separate validation of previous TEM/CD-AFM calibration. Excellent agreement was observed.

  17. A general analysis of calibrated BOLD methodology for measuring CMRO2 responses: comparison of a new approach with existing methods.

    PubMed

    Blockley, Nicholas P; Griffeth, Valerie E M; Buxton, Richard B

    2012-03-01

    The amplitude of the BOLD response to a stimulus is not only determined by changes in cerebral blood flow (CBF) and oxygen metabolism (CMRO(2)), but also by baseline physiological parameters such as haematocrit, oxygen extraction fraction (OEF) and blood volume. The calibrated BOLD approach aims to account for this physiological variation by performing an additional calibration scan. This calibration typically consists of a hypercapnia or hyperoxia respiratory challenge, although we propose that a measurement of the reversible transverse relaxation rate, R(2)', might also be used. A detailed model of the BOLD effect was used to simulate each of the calibration experiments, as well as the activation experiment, whilst varying a number of physiological parameters associated with the baseline state and response to activation. The effectiveness of the different calibration methods was considered by testing whether the BOLD response to activation scaled by the calibration parameter combined with the measured CBF provides sufficient information to reliably distinguish different levels of CMRO(2) response despite underlying physiological variability. In addition the effect of inaccuracies in the underlying assumptions of each technique were tested, e.g. isometabolism during hypercapnia. The three primary findings of the study were: 1) The new calibration method based on R(2)' worked reasonably well, although not as well as the ideal hypercapnia method; 2) The hyperoxia calibration method was significantly worse because baseline haematocrit and OEF must be assumed, and these physiological parameters have a significant effect on the measurements; and 3) the venous blood volume change with activation is an important confounding variable for all of the methods, with the hypercapnia method being the most robust when this is uncertain.

  18. A new systematic calibration method of ring laser gyroscope inertial navigation system

    NASA Astrophysics Data System (ADS)

    Wei, Guo; Gao, Chunfeng; Wang, Qi; Wang, Qun; Xiong, Zhenyu; Long, Xingwu

    2016-10-01

    Inertial navigation system has been the core component of both military and civil navigation systems. Before the INS is put into application, it is supposed to be calibrated in the laboratory in order to compensate repeatability error caused by manufacturing. Discrete calibration method cannot fulfill requirements of high-accurate calibration of the mechanically dithered ring laser gyroscope navigation system with shock absorbers. This paper has analyzed theories of error inspiration and separation in detail and presented a new systematic calibration method for ring laser gyroscope inertial navigation system. Error models and equations of calibrated Inertial Measurement Unit are given. Then proper rotation arrangement orders are depicted in order to establish the linear relationships between the change of velocity errors and calibrated parameter errors. Experiments have been set up to compare the systematic errors calculated by filtering calibration result with those obtained by discrete calibration result. The largest position error and velocity error of filtering calibration result are only 0.18 miles and 0.26m/s compared with 2 miles and 1.46m/s of discrete calibration result. These results have validated the new systematic calibration method and proved its importance for optimal design and accuracy improvement of calibration of mechanically dithered ring laser gyroscope inertial navigation system.

  19. Calibration with confidence: a principled method for panel assessment

    PubMed Central

    MacKay, R. S.; Low, R. J.; Parker, S.

    2017-01-01

    Frequently, a set of objects has to be evaluated by a panel of assessors, but not every object is assessed by every assessor. A problem facing such panels is how to take into account different standards among panel members and varying levels of confidence in their scores. Here, a mathematically based algorithm is developed to calibrate the scores of such assessors, addressing both of these issues. The algorithm is based on the connectivity of the graph of assessors and objects evaluated, incorporating declared confidences as weights on its edges. If the graph is sufficiently well connected, relative standards can be inferred by comparing how assessors rate objects they assess in common, weighted by the levels of confidence of each assessment. By removing these biases, ‘true’ values are inferred for all the objects. Reliability estimates for the resulting values are obtained. The algorithm is tested in two case studies: one by computer simulation and another based on realistic evaluation data. The process is compared to the simple averaging procedure in widespread use, and to Fisher's additive incomplete block analysis. It is anticipated that the algorithm will prove useful in a wide variety of situations such as evaluation of the quality of research submitted to national assessment exercises; appraisal of grant proposals submitted to funding panels; ranking of job applicants; and judgement of performances on degree courses wherein candidates can choose from lists of options. PMID:28386432

  20. Calibration method for the reference parameter in Fernald and Klett inversion combining Raman and Elastic return

    NASA Astrophysics Data System (ADS)

    Ji, Hongzhu; Chen, Siying; Zhang, Yinchao; Chen, He; Guo, Pan; Chen, Hao

    2017-02-01

    A calibration method is proposed to invert the extinction coefficient for Fernald and Klett inversion by using the particle backscattering coefficient inversed with Raman and Elastic return signals. The calibration method is analyzed theoretically and experimentally, the inversion accuracy can be improved by removing the dependence on reference altitudes and intervals in conventional calibration methods, which resulted from the introduction of backscattering coefficient with relatively higher accuracy obtained by Raman-Mie inversion method. The standard deviation of this new calibration method can be reduced by about 20×, compared to that of the conventional calibration methods of Fernald and Klett inversion. And, the more stable effective inversed range with this new calibration method can be obtained by removing the dimple phenomenon in clouds position.

  1. Improved camera calibration method based on perpendicularity compensation for binocular stereo vision measurement system.

    PubMed

    Jia, Zhenyuan; Yang, Jinghao; Liu, Wei; Wang, Fuji; Liu, Yang; Wang, Lingli; Fan, Chaonan; Zhao, Kai

    2015-06-15

    High-precision calibration of binocular vision systems plays an important role in accurate dimensional measurements. In this paper, an improved camera calibration method is proposed. First, an accurate intrinsic parameters calibration method based on active vision with perpendicularity compensation is developed. Compared to the previous work, this method eliminates the effect of non-perpendicularity of the camera motion on calibration accuracy. The principal point, scale factors, and distortion factors are calculated independently in this method, thereby allowing the strong coupling of these parameters to be eliminated. Second, an accurate global optimization method with only 5 images is presented. The results of calibration experiments show that the accuracy of the calibration method can reach 99.91%.

  2. Real Time Calibration Method for Signal Conditioning Amplifiers

    NASA Technical Reports Server (NTRS)

    Medelius, Pedro J. (Inventor); Mata, Carlos T. (Inventor); Eckhoff, Anthony (Inventor); Perotti, Jose (Inventor); Lucena, Angel (Inventor)

    2004-01-01

    A signal conditioning amplifier receives an input signal from an input such as a transducer. The signal is amplified and processed through an analog to digital converter and sent to a processor. The processor estimates the input signal provided by the transducer to the amplifier via a multiplexer. The estimated input signal is provided as a calibration voltage to the amplifier immediately following the receipt of the amplified input signal. The calibration voltage is amplified by the amplifier and provided to the processor as an amplified calibration voltage. The amplified calibration voltage is compared to the amplified input signal, and if a significant error exists, the gain and/or offset of the amplifier may be adjusted as necessary.

  3. Two-Step Camera Calibration Method Developed for Micro UAV'S

    NASA Astrophysics Data System (ADS)

    Gašparović, M.; Gajski, D.

    2016-06-01

    The development of unmanned aerial vehicles (UAVs) and continuous price reduction of unmanned systems attracted us to this research. Professional measuring systems are dozens of times more expensive and often heavier than "amateur", non-metric UAVs. For this reason, we tested the DJI Phantom 2 Vision Plus UAV. Phantom's smaller mass and velocity can develop less kinetic energy in relation to the professional measurement platforms, which makes it potentially less dangerous for use in populated areas. In this research, we wanted to investigate the ability of such non-metric UAV and find the procedures under which this kind of UAV may be used for the photogrammetric survey. It is important to emphasize that UAV is equipped with an ultra wide-angle camera with 14MP sensor. Calibration of such cameras is a complex process. In the research, a new two-step process is presented and developed, and the results are compared with standard one-step camera calibration procedure. Two-step process involves initially removed distortion on all images, and then uses these images in the phototriangulation with self-calibration. The paper presents statistical indicators which proved that the proposed two-step process is better and more accurate procedure for calibrating those types of cameras than standard one-step calibration. Also, we suggest two-step calibration process as the standard for ultra-wideangle cameras for unmanned aircraft.

  4. Research on method of geometry and spectral calibration of pushbroom dispersive hyperspectral imager

    NASA Astrophysics Data System (ADS)

    He, Zhiping; Shu, Rong; Wang, Jianyu

    2012-11-01

    Development and application of airborne and aerospace hyperspectral imager press for high precision geometry and spectral calibration of pixels of image cube. The research of geometry and spectral calibration of pushbroom hyperspectral imager, its target is giving the coordinate of angle field of view and center wavelength of each detect unit in focal plane detector of hyperspectral imager, and achieves the high precision, full field of view, full channel geometry and spectral calibration. It is importance for imaging quantitative and deep application of hyperspectal imager. The paper takes the geometry and spectral calibration of pushbroom dispersive hyperspectral imager as case study, and research on the constitution and analysis of imaging mathematical model. Aimed especially at grating-dispersive hyperspectral imaging, the specialty of the imaging mode and dispersive method has been concretely analyzed. Based on the analysis, the theory and feasible method of geometry and spectral calibration of dispersive hyperspectral imager is set up. The key technique has been solved is As follows: 1). the imaging mathematical model and feasible method of geometry and spectral calibration for full pixels of image cube has been set up, the feasibility of the calibration method has been analyzed. 2). the engineering model and method of the geometry and spectral calibration of pushbroom dispersive hyperspectral imager has been set up and the calibration equipment has been constructed, and the calibration precision has been analyzed.

  5. Optical geometry calibration method for free-form digital tomosynthesis

    NASA Astrophysics Data System (ADS)

    Chtcheprov, Pavel; Hartman, Allison; Shan, Jing; Lee, Yueh Z.; Zhou, Otto; Lu, Jianping

    2016-03-01

    Digital tomosynthesis is a type of limited angle tomography that allows 3D information to be reconstructed from a set of x-ray projection images taken at various angles using an x-ray tube, a mechanical arm to rotate the tube about the object, and a digital detector. Tomosynthesis reconstruction requires the precise location of the detector with respect to each x-ray source, forcing all current clinical tomosynthesis systems to use a physically coupled source and detector so the geometry is always known and is always the same. This limits the imaging geometries and its large size is impractical for mobile or field operations. To counter this, we have developed a free form tomosynthesis with a decoupled, free-moving source and detector that uses a novel optical method for accurate and real-time geometry calibration to allow for manual, hand-held tomosynthesis and even CT imaging. We accomplish this by using a camera, attached to the source, to track the motion of the source relative to the detector. Attached to the detector is an optical pattern and the image captured by the camera is then used to determine the relative camera/pattern position and orientation by analyzing the pattern distortion and calculating the source positions for each projection, necessary for 3D reconstruction. This allows for portable imaging in the field and also as an inexpensive upgrade to existing 2D systems, such as in developing countries, to provide 3D image data. Here we report the first feasibility demonstrations of free form digital tomosynthesis systems using the method.

  6. Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories.

    PubMed

    Hourdakis, Costas J; Büermann, Ludwig; Ciraj-Bjelac, Olivera; Csete, Istvan; Delis, Harry; Gomola, Igor; Persson, Linda; Novak, Leos; Petkov, Ivailo; Toroi, Paula

    2016-01-01

    A comparison of calibration results and procedures in terms of air kerma length product, PKL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the "partial irradiation method" is the preferred method for the pencil-type IC calibration in terms of PKL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement (>99%) for both dosimetry quantities (PKL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed.

  7. An additional study and implementation of tone calibrated technique of modulation

    NASA Technical Reports Server (NTRS)

    Rafferty, W.; Bechtel, L. K.; Lay, N. E.

    1985-01-01

    The Tone Calibrated Technique (TCT) was shown to be theoretically free from an error floor, and is only limited, in practice, by implementation constraints. The concept of the TCT transmission scheme along with a baseband implementation of a suitable demodulator is introduced. Two techniques for the generation of the TCT signal are considered: a Manchester source encoding scheme (MTCT) and a subcarrier based technique (STCT). The results are summarized for the TCT link computer simulation. The hardware implementation of the MTCT system is addressed and the digital signal processing design considerations involved in satisfying the modulator/demodulator requirements are outlined. The program findings are discussed and future direction are suggested based on conclusions made regarding the suitability of the TCT system for the transmission channel presently under consideration.

  8. A Quantitative Comparison of Calibration Methods for RGB-D Sensors Using Different Technologies

    PubMed Central

    Villena-Martínez, Víctor; Fuster-Guilló, Andrés; Azorín-López, Jorge; Saval-Calvo, Marcelo; Mora-Pascual, Jeronimo; Garcia-Rodriguez, Jose; Garcia-Garcia, Alberto

    2017-01-01

    RGB-D (Red Green Blue and Depth) sensors are devices that can provide color and depth information from a scene at the same time. Recently, they have been widely used in many solutions due to their commercial growth from the entertainment market to many diverse areas (e.g., robotics, CAD, etc.). In the research community, these devices have had good uptake due to their acceptable level of accuracy for many applications and their low cost, but in some cases, they work at the limit of their sensitivity, near to the minimum feature size that can be perceived. For this reason, calibration processes are critical in order to increase their accuracy and enable them to meet the requirements of such kinds of applications. To the best of our knowledge, there is not a comparative study of calibration algorithms evaluating its results in multiple RGB-D sensors. Specifically, in this paper, a comparison of the three most used calibration methods have been applied to three different RGB-D sensors based on structured light and time-of-flight. The comparison of methods has been carried out by a set of experiments to evaluate the accuracy of depth measurements. Additionally, an object reconstruction application has been used as example of an application for which the sensor works at the limit of its sensitivity. The obtained results of reconstruction have been evaluated through visual inspection and quantitative measurements. PMID:28134826

  9. A method of diameter measurement for spur gear based on camera calibration

    NASA Astrophysics Data System (ADS)

    Wu, Ziyue; Geng, Jinfeng; Xu, Zhe

    2012-04-01

    The camera calibration is the basis of putting the computer vision technology into practice. This paper proposes a new method based on camera calibration for diameter measurement of gear, and analyses the error from calibration and measurement. Diameter values are gained by this method, which firstly gets the intrinsic parameters and the extrinsic parameters by camera calibration, then transforms the feature points in image coordinate extracted from the image plane of gear to the 3D world coordinate, lastly computes distance between the features points. The experiment results demonstrate that the method is simple and quick, and easy to implement, highly precise, and rarely limited to the size of target.

  10. Model Robust Calibration: Method and Application to Electronically-Scanned Pressure Transducers

    NASA Technical Reports Server (NTRS)

    Walker, Eric L.; Starnes, B. Alden; Birch, Jeffery B.; Mays, James E.

    2010-01-01

    This article presents the application of a recently developed statistical regression method to the controlled instrument calibration problem. The statistical method of Model Robust Regression (MRR), developed by Mays, Birch, and Starnes, is shown to improve instrument calibration by reducing the reliance of the calibration on a predetermined parametric (e.g. polynomial, exponential, logarithmic) model. This is accomplished by allowing fits from the predetermined parametric model to be augmented by a certain portion of a fit to the residuals from the initial regression using a nonparametric (locally parametric) regression technique. The method is demonstrated for the absolute scale calibration of silicon-based pressure transducers.

  11. Indoor Calibration for Stereoscopic Camera STC, A New Method

    NASA Astrophysics Data System (ADS)

    Simioni, E.; Re, C.; Da Deppo, V.; Naletto, G.; Borrelli, D.; Dami, M.; Ficai Veltroni, I.; Cremonese, G.

    2014-10-01

    In the framework of the ESA-JAXA BepiColombo mission to Mercury, the global mapping of the planet will be performed by the on-board Stereo Camera (STC), part of the SIMBIO-SYS suite [1]. In this paper we propose a new technique for the validation of the 3D reconstruction of planetary surface from images acquired with a stereo camera. STC will provide a three-dimensional reconstruction of Mercury surface. The generation of a DTM of the observed features is based on the processing of the acquired images and on the knowledge of the intrinsic and extrinsic parameters of the optical system. The new stereo concept developed for STC needs a pre-flight verification of the actual capabilities to obtain elevation information from stereo couples: for this, a stereo validation setup to get an indoor reproduction of the flight observing condition of the instrument would give a much greater confidence to the developed instrument design. STC is the first stereo satellite camera with two optical channels converging in a unique sensor. Its optical model is based on a brand new concept to minimize mass and volume and to allow push-frame imaging. This model imposed to define a new calibration pipeline to test the reconstruction method in a controlled ambient. An ad-hoc indoor set-up has been realized for validating the instrument designed to operate in deep space, i.e. in-flight STC will have to deal with source/target essentially placed at infinity. This auxiliary indoor setup permits on one side to rescale the stereo reconstruction problem from the operative distance in-flight of 400 km to almost 1 meter in lab; on the other side it allows to replicate different viewing angles for the considered targets. Neglecting for sake of simplicity the Mercury curvature, the STC observing geometry of the same portion of the planet surface at periherm corresponds to a rotation of the spacecraft (SC) around the observed target by twice the 20° separation of each channel with respect to nadir

  12. Calibration of three rainfall simulators with automatic measurement methods

    NASA Astrophysics Data System (ADS)

    Roldan, Margarita

    2010-05-01

    CALIBRATION OF THREE RAINFALL SIMULATORS WITH AUTOMATIC MEASUREMENT METHODS M. Roldán (1), I. Martín (2), F. Martín (2), S. de Alba(3), M. Alcázar(3), F.I. Cermeño(3) 1 Grupo de Investigación Ecología y Gestión Forestal Sostenible. ECOGESFOR-Universidad Politécnica de Madrid. E.U.I.T. Forestal. Avda. Ramiro de Maeztu s/n. Ciudad Universitaria. 28040 Madrid. margarita.roldan@upm.es 2 E.U.I.T. Forestal. Avda. Ramiro de Maeztu s/n. Ciudad Universitaria. 28040 Madrid. 3 Facultad de Ciencias Geológicas. Universidad Complutense de Madrid. Ciudad Universitaria s/n. 28040 Madrid The rainfall erosivity is the potential ability of rain to cause erosion. It is function of the physical characteristics of rainfall (Hudson, 1971). Most expressions describing erosivity are related to kinetic energy or momentum and so with drop mass or size and fall velocity. Therefore, research on factors determining erosivity leds to the necessity to study the relation between fall height and fall velocity for different drop sizes, generated in a rainfall simulator (Epema G.F.and Riezebos H.Th, 1983) Rainfall simulators are one of the most used tools for erosion studies and are used to determine fall velocity and drop size. Rainfall simulators allow repeated and multiple measurements The main reason for use of rainfall simulation as a research tool is to reproduce in a controlled way the behaviour expected in the natural environment. But in many occasions when simulated rain is used in order to compare it with natural rain, there is a lack of correspondence between natural and simulated rain and this can introduce some doubt about validity of data because the characteristics of natural rain are not adequately represented in rainfall simulation research (Dunkerley D., 2008). Many times the rainfall simulations have high rain rates and they do not resemble natural rain events and these measures are not comparables. And besides the intensity is related to the kinetic energy which

  13. Two methods for absolute calibration of dynamic pressure transducers

    NASA Astrophysics Data System (ADS)

    Swift, G. W.; Migliori, A.; Garrett, S. L.; Wheatley, J. C.

    1982-12-01

    Two techniques are described for absolute calibration of a dynamic pressure transducer from 0 to 400 Hz in 1-MPa helium gas. One technique is based on a comparison to a mercury manometer; the other is based on the principle of reciprocity. The two techniques agree within the instrumental uncertainties of 1%.

  14. Methods to calibrate the absolute receive sensitivity of single-element, focused transducers.

    PubMed

    Rich, Kyle T; Mast, T Douglas

    2015-09-01

    Absolute pressure measurements of acoustic emissions by single-element, focused passive cavitation detectors would be facilitated by improved wideband receive calibration techniques. Here, calibration methods were developed to characterize the absolute, frequency-dependent receive sensitivity of a spherically focused, single-element transducer using pulse-echo and pitch-catch techniques. Validation of these calibration methods on a focused receiver were made by generating a pulse from a small diameter source at the focus of the transducer and comparing the absolute pressure measured by a calibrated hydrophone to that of the focused transducer using the receive sensitivities determined here.

  15. System and Method for the Calibration of a Hydrophone Line Array

    DTIC Science & Technology

    2014-09-25

    b . ABSTRACT c. THIS PAGE 19b. TELEPHONE NUMBER (include area code) Standard Form 298 (Re . 8-98) v Prescribed by ANSI Std. Z39.18 05...2, 2013 by the inventors, Steven E. Crocker, Daniel C. Casimiro, Robert F . Cutler, Ronald R. Smalley and entitled “METHOD FOR THE CALIBRATION OF A...practice is often referred to as a relative calibration. [0008] In Skinner (“Place Calibration of Sonar Receive Array”; United States Patent No

  16. Calibration method of tilt and azimuth angles for alignment of TEM tomographic tilt series

    NASA Astrophysics Data System (ADS)

    Hayashida, Misa; Terauchi, Shinya; Fujimoto, Toshiyuki

    2011-10-01

    This paper describes the calibration method of the tilt and azimuth angles of specimen using a digital protractor and a laser autocollimator for alignment of electron tomography. It also suggests an easy method to check whether the specimen is tilted by 180.0°, and whether the azimuth angle is 0.0°; the method involves the use of two images of a rod-shaped specimen collected before and after a 180.0° tilt. The method is based on the assumption that these images are symmetric about the tilt axis when the azimuth angle is 0.0°. In addition, we used an experiment to demonstrate the effect of the incorrect angles on reconstructed images and simulated the image quality against distance away from tilt axis.

  17. Shortwave Radiometer Calibration Methods Comparison and Resulting Solar Irradiance Measurement Differences: A User Perspective

    SciTech Connect

    Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Reda, Ibrahim; Robinson, Justin

    2016-11-21

    Banks financing solar energy projects require assurance that these systems will produce the energy predicted. Furthermore, utility planners and grid system operators need to understand the impact of the variable solar resource on solar energy conversion system performance. Accurate solar radiation data sets reduce the expense associated with mitigating performance risk and assist in understanding the impacts of solar resource variability. The accuracy of solar radiation measured by radiometers depends on the instrument performance specification, installation method, calibration procedure, measurement conditions, maintenance practices, location, and environmental conditions. This study addresses the effect of different calibration methods provided by radiometric calibration service providers, such as NREL and manufacturers of radiometers, on the resulting calibration responsivity. Some of these radiometers are calibrated indoors and some outdoors. To establish or understand the differences in calibration methodology, we processed and analyzed field-measured data from these radiometers. This study investigates calibration responsivities provided by NREL's broadband outdoor radiometer calibration (BORCAL) and a few prominent manufacturers. The BORCAL method provides the outdoor calibration responsivity of pyranometers and pyrheliometers at 45 degree solar zenith angle, and as a function of solar zenith angle determined by clear-sky comparisons with reference irradiance. The BORCAL method also employs a thermal offset correction to the calibration responsivity of single-black thermopile detectors used in pyranometers. Indoor calibrations of radiometers by their manufacturers are performed using a stable artificial light source in a side-by-side comparison between the test radiometer under calibration and a reference radiometer of the same type. In both methods, the reference radiometer calibrations are traceable to the World Radiometric Reference (WRR). These different

  18. A self-calibration method for tri-axis rotational inertial navigation system

    NASA Astrophysics Data System (ADS)

    Gao, Pengyu; Li, Kui; Wang, Lei; Liu, Zengjun

    2016-11-01

    The navigation accuracy of the rotational inertial navigation system (RINS) could be greatly improved by periodically rotating the inertial measurement unit (IMU) with gimbals. However, error parameters in RINS should be effectively calibrated and compensated. In this paper, a self-calibration method is proposed for tri-axis RINS using attitude errors and velocity errors as measurements. The proposed calibration scheme is designed as three separate steps, and a certain gimbal rotates continuously in each step. All the error parameters in the RINS are calibrated when the whole scheme finishes. The separate calibration steps reduce the correlations between error parameters, and the observability of errors in this method is clear to demonstrate according to the relations between navigation errors and error parameters when gimbals rotate. Each calibration step only lasts 12 min, thus gyro drifts and accelerometers biases could be regarded as constant. The proposed calibration scheme is tested in both simulation and actual tri-axis RINS, and simulation and experimental results show that all 23 error parameters could be well estimated in tri-axis RINS. A long-term vehicle navigation experiment results show that after calibration and compensation, the navigation performance has doubled approximately, and the velocity accuracy is less than 2 m s-1 while the position accuracy is less than 1500 m, fully illustrating the significance of the proposed self-calibration method in improving the navigation performance of RINS.

  19. Method and apparatus for calibrating a tiled display

    NASA Technical Reports Server (NTRS)

    Johnson, Michael J. (Inventor); Chen, Chung-Jen (Inventor); Chandrasekhar, Rajesh (Inventor)

    2001-01-01

    A display system that can be calibrated and re-calibrated with a minimal amount of manual intervention. To accomplish this, one or more cameras are provided to capture an image of the display screen. The resulting captured image is processed to identify any non-desirable characteristics, including visible artifacts such as seams, bands, rings, etc. Once the non-desirable characteristics are identified, an appropriate transformation function is determined. The transformation function is used to pre-warp the input video signal that is provided to the display such that the non-desirable characteristics are reduced or eliminated from the display. The transformation function preferably compensates for spatial non-uniformity, color non-uniformity, luminance non-uniformity, and other visible artifacts.

  20. Camera calibration method of binocular stereo vision based on OpenCV

    NASA Astrophysics Data System (ADS)

    Zhong, Wanzhen; Dong, Xiaona

    2015-10-01

    Camera calibration, an important part of the binocular stereo vision research, is the essential foundation of 3D reconstruction of the spatial object. In this paper, the camera calibration method based on OpenCV (open source computer vision library) is submitted to make the process better as a result of obtaining higher precision and efficiency. First, the camera model in OpenCV and an algorithm of camera calibration are presented, especially considering the influence of camera lens radial distortion and decentering distortion. Then, camera calibration procedure is designed to compute those parameters of camera and calculate calibration errors. High-accurate profile extraction algorithm and a checkboard with 48 corners have also been used in this part. Finally, results of calibration program are presented, demonstrating the high efficiency and accuracy of the proposed approach. The results can reach the requirement of robot binocular stereo vision.

  1. A method to calibrate a camera using perpendicularity of 2D lines in the target observations

    PubMed Central

    Xu, Guan; Zheng, Anqi; Li, Xiaotao; Su, Jian

    2016-01-01

    Camera calibration based on point features leads the main trends in vision-based measurement systems for both fundamental researches and potential applications. However, the calibration results tend to be affected by the precision of the feature point extraction in the camera images. As the point features are noise sensitive, line features are more appropriate to provide a stable calibration due to the noise immunity of line features. We propose a calibration method using the perpendicularity of the lines on a 2D target. The objective function of the camera internal parameters is theoretically constructed by the reverse projections of the image lines on a 2D target in the world coordinate system. We experimentally explore the performances of the perpendicularity method and compare them with the point feature methods at different distances. By the perpendicularity and the noise immunity of the lines, our work achieves a relatively higher calibration precision. PMID:27713566

  2. A New Method of Theodolite Calibration Based on Image Processing Technology

    NASA Astrophysics Data System (ADS)

    Zou, Hui-Hui; Wu, Hong-Bing; Chen, Di

    Aiming at improving the theodolite calibration method for space tracking ship, a calibration device which consists of hardware and software is designed in this paper. Hereinto, the hardware part is a set of optical acquisition system that includes CCD, lens and 0.2" collimator, while the software part contains image acquisition module, image processing module, data processing module and interface display module. During the calibration process, the new methods of image denoising and image character extraction are applied to improve the precision of image measure. The result of the experiment shows that the calibration criteria of the theodolite errors was met by applying the image processing technology of the theodolite calibration device, it is more accurate than the manual reading method under the same situation in dock.

  3. A method to calibrate a camera using perpendicularity of 2D lines in the target observations

    NASA Astrophysics Data System (ADS)

    Xu, Guan; Zheng, Anqi; Li, Xiaotao; Su, Jian

    2016-10-01

    Camera calibration based on point features leads the main trends in vision-based measurement systems for both fundamental researches and potential applications. However, the calibration results tend to be affected by the precision of the feature point extraction in the camera images. As the point features are noise sensitive, line features are more appropriate to provide a stable calibration due to the noise immunity of line features. We propose a calibration method using the perpendicularity of the lines on a 2D target. The objective function of the camera internal parameters is theoretically constructed by the reverse projections of the image lines on a 2D target in the world coordinate system. We experimentally explore the performances of the perpendicularity method and compare them with the point feature methods at different distances. By the perpendicularity and the noise immunity of the lines, our work achieves a relatively higher calibration precision.

  4. Improvement of Accuracy in Environmental Dosimetry by TLD Cards Using Three-dimensional Calibration Method

    PubMed Central

    HosseiniAliabadi, S. J.; Hosseini Pooya, S. M.; Afarideh, H.; Mianji, F.

    2015-01-01

    Introduction The angular dependency of response for TLD cards may cause deviation from its true value on the results of environmental dosimetry, since TLDs may be exposed to radiation at different angles of incidence from the surrounding area. Objective A 3D setting of TLD cards has been calibrated isotropically in a standard radiation field to evaluate the improvement of the accuracy of measurement for environmental dosimetry. Method Three personal TLD cards were rectangularly placed in a cylindrical holder, and calibrated using 1D and 3D calibration methods. Then, the dosimeter has been used simultaneously with a reference instrument in a real radiation field measuring the accumulated dose within a time interval. Result The results show that the accuracy of measurement has been improved by 6.5% using 3D calibration factor in comparison with that of normal 1D calibration method. Conclusion This system can be utilized in large scale environmental monitoring with a higher accuracy. PMID:26157729

  5. Research on global calibration method for multi-camera visual measurement system

    NASA Astrophysics Data System (ADS)

    Huang, Dongzhao; Zhao, Qiancheng; Ou, Yun; Yang, Tianlong

    2016-01-01

    A key technology of multi-camera visual measurement system is global calibration. The global calibration methods existed either has disadvantages such as high cost and complicated operation, or has limited application areas only for visual system based on stereo vision. A convenient global calibration method for multi-camera system based on two planar targets is proposed in this paper, and the pose relation between the two targets can be unknown. This method is not only suitable for system based on stereo vision, but also for system based on monocular vision. It has been used in the factory-calibration of four-wheel aligner consisted of 2 or 4 cameras, and calibration precision meet the requirement.

  6. Anisotropic phantom to calibrate high-q diffusion MRI methods

    NASA Astrophysics Data System (ADS)

    Komlosh, M. E.; Benjamini, D.; Barnett, A. S.; Schram, V.; Horkay, F.; Avram, A. V.; Basser, P. J.

    2017-02-01

    A silicon oil-filled glass capillary array is proposed as an anisotropic diffusion MRI phantom. Together with a computational/theoretical pipeline these provide a gold standard for calibrating and validating high-q diffusion MRI experiments. The phantom was used to test high angular resolution diffusion imaging (HARDI) and double pulsed-field gradient (d-PFG) MRI acquisition schemes. MRI-based predictions of microcapillary diameter using both acquisition schemes were compared with results from optical microscopy. This phantom design can be used for quality control and quality assurance purposes and for testing and validating proposed microstructure imaging experiments and the processing pipelines used to analyze them.

  7. Performance analysis for time-frequency MUSIC algorithm in presence of both additive noise and array calibration errors

    NASA Astrophysics Data System (ADS)

    Khodja, Mohamed; Belouchrani, Adel; Abed-Meraim, Karim

    2012-12-01

    This article deals with the application of Spatial Time-Frequency Distribution (STFD) to the direction finding problem using the Multiple Signal Classification (MUSIC)algorithm. A comparative performance analysis is performed for the method under consideration with respect to that using data covariance matrix when the received array signals are subject to calibration errors in a non-stationary environment. An unified analytical expression of the Direction Of Arrival (DOA) error estimation is derived for both methods. Numerical results show the effect of the parameters intervening in the derived expression on the algorithm performance. It is particularly observed that for low Signal to Noise Ratio (SNR) and high Signal to sensor Perturbation Ratio (SPR) the STFD method gives better performance, while for high SNR and for the same SPR both methods give similar performance.

  8. Design of a Two-Step Calibration Method of Kinematic Parameters for Serial Robots

    NASA Astrophysics Data System (ADS)

    WANG, Wei; WANG, Lei; YUN, Chao

    2017-03-01

    Serial robots are used to handle workpieces with large dimensions, and calibrating kinematic parameters is one of the most efficient ways to upgrade their accuracy. Many models are set up to investigate how many kinematic parameters can be identified to meet the minimal principle, but the base frame and the kinematic parameter are indistinctly calibrated in a one-step way. A two-step method of calibrating kinematic parameters is proposed to improve the accuracy of the robot's base frame and kinematic parameters. The forward kinematics described with respect to the measuring coordinate frame are established based on the product-of-exponential (POE) formula. In the first step the robot's base coordinate frame is calibrated by the unit quaternion form. The errors of both the robot's reference configuration and the base coordinate frame's pose are equivalently transformed to the zero-position errors of the robot's joints. The simplified model of the robot's positioning error is established in second-power explicit expressions. Then the identification model is finished by the least square method, requiring measuring position coordinates only. The complete subtasks of calibrating the robot's 39 kinematic parameters are finished in the second step. It's proved by a group of calibration experiments that by the proposed two-step calibration method the average absolute accuracy of industrial robots is updated to 0.23 mm. This paper presents that the robot's base frame should be calibrated before its kinematic parameters in order to upgrade its absolute positioning accuracy.

  9. Spectral responsivity calibrations of two types of pyroelectric radiometers using three different methods

    NASA Astrophysics Data System (ADS)

    Zeng, J.; Eppeldauer, G. P.; Hanssen, L. M.; Podobedov, V. B.

    2012-06-01

    Spectral responsivity calibrations of two different types of pyroelectric radiometers have been made in the infrared region up to 14 μm in power mode using three different calibration facilities at NIST. One pyroelectric radiometer is a temperature-controlled low noise-equivalent-power (NEP) single-element pyroelectric radiometer with an active area of 5 mm in diameter. The other radiometer is a prototype using the same type of pyroeletric detector with dome-input optics, which was designed to increase absorptance and to minimize spectral structures to obtain a constant spectral responsivity. Three calibration facilities at NIST were used to conduct direct and indirect responsivity calibrations tied to absolute scales in the infrared spectral regime. We report the calibration results for the single-element pyroelectric radiometer using a new Infrared Spectral Comparator Facility (IRSCF) for direct calibration. Also, a combined method using the Fourier Transform Infrared Spectrophotometry (FTIS) facility and single wavelength laser tie-points are described to calibrated standard detectors with an indirect approach. For the dome-input pyroelectric radiometer, the results obtained from another direct calibration method using a circular variable filter (CVF) spectrometer and the FTIS are also presented. The inter-comparison of different calibration methods enables us to improve the responsivity uncertainty performed by the different facilities. For both radiometers, consistent results of the spectral power responsivity have been obtained applying different methods from 1.5 μm to 14 μm with responsivity uncertainties between 1 % and 2 % (k = 2). Relevant characterization results, such as spatial uniformity, linearity, and angular dependence of responsivity, are shown. Validation of the spectral responsivity calibrations, uncertainty sources, and improvements for each method will also be discussed.

  10. A novel calibration method for an infrared thermography system applied to heat transfer experiments

    NASA Astrophysics Data System (ADS)

    Ochs, M.; Horbach, T.; Schulz, A.; Koch, R.; Bauer, H.-J.

    2009-07-01

    In heat transfer measurements with highly non-uniform wall heat fluxes, high spatial resolution of wall temperatures is required to fully capture the complex thermal situation. Infrared thermography systems provide that spatial resolution. To meet the thermal accuracy, they are usually calibrated in situ using thermocouples embedded in the test surface, which have to cover the complete temperature range of interest. However, thermocouples which are placed in regions of high temperature and heat flux gradients often cannot be used for the calibration and the overall accuracy of the calibration decreases significantly. Therefore, in the present work a novel in situ calibration method is presented which does not require thermocouples over the complete surface temperature range. The number of free parameters of the calibration function is reduced by an optimized insensitivity of the system with respect to changes in operating conditions. Reference measurements demonstrate the advantages of the new method.

  11. In Search of Easy-to-Use Methods for Calibrating ADCP's for Velocity and Discharge Measurements

    USGS Publications Warehouse

    Oberg, K.; ,

    2002-01-01

    A cost-effective procedure for calibrating acoustic Doppler current profilers (ADCP) in the field was presented. The advantages and disadvantages of various methods which are used for calibrating ADCP were discussed. The proposed method requires the use of differential global positioning system (DGPS) with sub-meter accuracy and standard software for collecting ADCP data. The method involves traversing a long (400-800 meter) course at a constant compass heading and speed, while collecting simultaneous DGPS and ADCP data.

  12. Self-calibration method for rotating laser positioning system using interscanning technology and ultrasonic ranging.

    PubMed

    Wu, Jun; Yu, Zhijing; Zhuge, Jingchang

    2016-04-01

    A rotating laser positioning system (RLPS) is an efficient measurement method for large-scale metrology. Due to multiple transmitter stations, which consist of a measurement network, the position relationship of these stations must be first calibrated. However, with such auxiliary devices such as a laser tracker, scale bar, and complex calibration process, the traditional calibration methods greatly reduce the measurement efficiency. This paper proposes a self-calibration method for RLPS, which can automatically obtain the position relationship. The method is implemented through interscanning technology by using a calibration bar mounted on the transmitter station. Each bar is composed of three RLPS receivers and one ultrasonic sensor whose coordinates are known in advance. The calibration algorithm is mainly based on multiplane and distance constraints and is introduced in detail through a two-station mathematical model. The repeated experiments demonstrate that the coordinate measurement uncertainty of spatial points by using this method is about 0.1 mm, and the accuracy experiments show that the average coordinate measurement deviation is about 0.3 mm compared with a laser tracker. The accuracy can meet the requirements of most applications, while the calibration efficiency is significantly improved.

  13. Effect of calibration on dispersion of glycohemoglobin values determined by 111 laboratories using 21 methods.

    PubMed

    Weykamp, C W; Penders, T J; Muskiet, F A; van der Slik, W

    1994-01-01

    One hundred eleven laboratories, using 21 different methods based on five different principles, determined glycohemoglobin (GHb) percentages in two identical series of six lyophilized hemolysates and three similarly processed calibrators, distributed 3 months apart. To assign GHb percentages to calibrators, we used HbA1c results from nine participants who used the Bio-Rad Diamat high-performance liquid chromatographic method. Three-point calibration with assigned values improved mean intralaboratory variation (CV) from 6.6% to 3.5%. For samples with low (5.5%) and high (14.1%) GHb percentages, respectively, calibration decreased interlaboratory variation per method (from 10% to 4% and from 6% to 3%), inter-method variation (from 18% to 4% and from 16% to 3%), and overall interlaboratory variation (from 25% to 7% and from 15% to 4%). Without calibration, 71% of the laboratories did not meet the clinically desirable intralaboratory CV of 3.5%; calibration reduced this proportion to 39%. We conclude that, irrespective of the analytical method used, calibration greatly reduces all sources of GHb variation.

  14. Calibration method for lumbosacral dimensions in wearable sensor system of lumbar alignment.

    PubMed

    Tsuchiya, Yoshio; Kusaka, Takashi; Tanaka, Takayuki; Matsuo, Yoshikazu; Oda, Makoto; Sasaki, Tsukasa; Kamishima, Tamotsu; Yamanaka, Masanori

    2015-01-01

    Anteflexion of the spine is essential for many physical activities in everyday life. However, this motion places the lumbar disks under heavy load due to changes in the shape of the lumbar spine and can lead to low back pain. With the aim of reducing low back pain, here we developed a wearable sensor system that can estimate lumbosacral alignment and lumbar load by measuring the shape of the lumbar skin when the lumbosacral alignment changes. In addition, we used this system to measure the parameters of anteflexion and studied the change in dimensions of the lumbar spine from changes in posture. By determining the dimensions of the lumbosacral spine on an X-ray image, a lumbosacral dimensions calibration method based on body surface area and height was developed. By using this method, lumbosacral alignment and lumbar load could be accurately estimated using the wearable sensor system.

  15. Alternate calibration method of radiochromic EBT3 film for quality assurance verification of clinical radiotherapy treatments

    NASA Astrophysics Data System (ADS)

    Park, Soah; Kang, Sei-Kwon; Cheong, Kwang-Ho; Hwang, Taejin; Yoon, Jai-Woong; Koo, Taeryool; Han, Tae Jin; Kim, Haeyoung; Lee, Me Yeon; Bae, Hoonsik; Kim, Kyoung Ju

    2016-07-01

    EBT3 film is utilized as a dosimetry quality assurance tool for the verification of clinical radiotherapy treatments. In this work, we suggest a percentage-depth-dose (PDD) calibration method that can calibrate several EBT3 film pieces together at different dose levels because photon beams provide different dose levels at different depths along the axis of the beam. We investigated the feasibility of the film PDD calibration method based on PDD data and compared the results those from the traditional film calibration method. Photon beams at 6 MV were delivered to EBT3 film pieces for both calibration methods. For the PDD-based calibration, the film pieces were placed on solid phantoms at the depth of maximum dose (dmax) and at depths of 3, 5, 8, 12, 17, and 22 cm, and a photon beam was delivered twice, at 100 cGy and 400 cGy, to extend the calibration dose range under the same conditions. Fourteen film pieces, to maintain their consistency, were irradiated at doses ranging from approximately 30 to 400 cGy for both film calibrations. The film pieces were located at the center position on the scan bed of an Epson 1680 flatbed scanner in the parallel direction. Intensity-modulated radiation therapy (IMRT) plans were created, and their dose distributions were delivered to the film. The dose distributions for the traditional method and those for the PDD-based calibration method were evaluated using a Gamma analysis. The PDD dose values using a CC13 ion chamber and those obtained by using a FC65-G Farmer chamber and measured at the depth of interest produced very similar results. With the objective test criterion of a 1% dosage agreement at 1 mm, the passing rates for the four cases of the three IMRT plans were essentially identical. The traditional and the PDD-based calibrations provided similar plan verification results. We also describe another alternative for calibrating EBT3 films, i.e., a PDD-based calibration method that provides an easy and time-saving approach

  16. A new method for automated dynamic calibration of tipping-bucket rain gauges

    USGS Publications Warehouse

    Humphrey, M.D.; Istok, J.D.; Lee, J.Y.; Hevesi, J.A.; Flint, A.L.

    1997-01-01

    Existing methods for dynamic calibration of tipping-bucket rain gauges (TBRs) can be time consuming and labor intensive. A new automated dynamic calibration system has been developed to calibrate TBRs with minimal effort. The system consists of a programmable pump, datalogger, digital balance, and computer. Calibration is performed in two steps: 1) pump calibration and 2) rain gauge calibration. Pump calibration ensures precise control of water flow rates delivered to the rain gauge funnel; rain gauge calibration ensures precise conversion of bucket tip times to actual rainfall rates. Calibration of the pump and one rain gauge for 10 selected pump rates typically requires about 8 h. Data files generated during rain gauge calibration are used to compute rainfall intensities and amounts from a record of bucket tip times collected in the field. The system was tested using 5 types of commercial TBRs (15.2-, 20.3-, and 30.5-cm diameters; 0.1-, 0.2-, and 1.0-mm resolutions) and using 14 TBRs of a single type (20.3-cm diameter; 0.1-mm resolution). Ten pump rates ranging from 3 to 154 mL min-1 were used to calibrate the TBRs and represented rainfall rates between 6 and 254 mm h-1 depending on the rain gauge diameter. All pump calibration results were very linear with R2 values greater than 0.99. All rain gauges exhibited large nonlinear underestimation errors (between 5% and 29%) that decreased with increasing rain gauge resolution and increased with increasing rainfall rate, especially for rates greater than 50 mm h-1. Calibration curves of bucket tip time against the reciprocal of the true pump rate for all rain gauges also were linear with R2 values of 0.99. Calibration data for the 14 rain gauges of the same type were very similar, as indicated by slope values that were within 14% of each other and ranged from about 367 to 417 s mm h-1. The developed system can calibrate TBRs efficiently, accurately, and virtually unattended and could be modified for use with other

  17. a Method for Self-Calibration in Satellite with High Precision of Space Linear Array Camera

    NASA Astrophysics Data System (ADS)

    Liu, Wei; Qian, Fangming; Miao, Yuzhe; Wang, Rongjian

    2016-06-01

    At present, the on-orbit calibration of the geometric parameters of a space surveying camera is usually processed by data from a ground calibration field after capturing the images. The entire process is very complicated and lengthy and cannot monitor and calibrate the geometric parameters in real time. On the basis of a large number of on-orbit calibrations, we found that owing to the influence of many factors, e.g., weather, it is often difficult to capture images of the ground calibration field. Thus, regular calibration using field data cannot be ensured. This article proposes a real time self-calibration method for a space linear array camera on a satellite using the optical auto collimation principle. A collimating light source and small matrix array CCD devices are installed inside the load system of the satellite; these use the same light path as the linear array camera. We can extract the location changes of the cross marks in the matrix array CCD to determine the real-time variations in the focal length and angle parameters of the linear array camera. The on-orbit status of the camera is rapidly obtained using this method. On one hand, the camera's change regulation can be mastered accurately and the camera's attitude can be adjusted in a timely manner to ensure optimal photography; in contrast, self-calibration of the camera aboard the satellite can be realized quickly, which improves the efficiency and reliability of photogrammetric processing.

  18. An Accurate Calibration Method Based on Velocity in a Rotational Inertial Navigation System.

    PubMed

    Zhang, Qian; Wang, Lei; Liu, Zengjun; Feng, Peide

    2015-07-28

    Rotation modulation is an effective method to enhance the accuracy of an inertial navigation system (INS) by modulating the gyroscope drifts and accelerometer bias errors into periodically varying components. The typical RINS drives the inertial measurement unit (IMU) rotation along the vertical axis and the horizontal sensors' errors are modulated, however, the azimuth angle error is closely related to vertical gyro drift, and the vertical gyro drift also should be modulated effectively. In this paper, a new rotation strategy in a dual-axis rotational INS (RINS) is proposed and the drifts of three gyros could be modulated, respectively. Experimental results from a real dual-axis RINS demonstrate that the maximum azimuth angle error is decreased from 0.04° to less than 0.01° during 1 h. Most importantly, the changing of rotation strategy leads to some additional errors in the velocity which is unacceptable in a high-precision INS. Then the paper studies the basic reason underlying horizontal velocity errors in detail and a relevant new calibration method is designed. Experimental results show that after calibration and compensation, the fluctuation and stages in the velocity curve disappear and velocity precision is improved.

  19. An Accurate Calibration Method Based on Velocity in a Rotational Inertial Navigation System

    PubMed Central

    Zhang, Qian; Wang, Lei; Liu, Zengjun; Feng, Peide

    2015-01-01

    Rotation modulation is an effective method to enhance the accuracy of an inertial navigation system (INS) by modulating the gyroscope drifts and accelerometer bias errors into periodically varying components. The typical RINS drives the inertial measurement unit (IMU) rotation along the vertical axis and the horizontal sensors’ errors are modulated, however, the azimuth angle error is closely related to vertical gyro drift, and the vertical gyro drift also should be modulated effectively. In this paper, a new rotation strategy in a dual-axis rotational INS (RINS) is proposed and the drifts of three gyros could be modulated, respectively. Experimental results from a real dual-axis RINS demonstrate that the maximum azimuth angle error is decreased from 0.04° to less than 0.01° during 1 h. Most importantly, the changing of rotation strategy leads to some additional errors in the velocity which is unacceptable in a high-precision INS. Then the paper studies the basic reason underlying horizontal velocity errors in detail and a relevant new calibration method is designed. Experimental results show that after calibration and compensation, the fluctuation and stages in the velocity curve disappear and velocity precision is improved. PMID:26225983

  20. Calibration methods and tools for KM3NeT

    NASA Astrophysics Data System (ADS)

    Kulikovskiy, Vladimir

    2016-04-01

    The KM3NeT detectors, ARCA and ORCA, composed of several thousands digital optical modules, are in the process of their realization in the Mediterranean Sea. Each optical module contains 31 3-inch photomultipliers. Readout of the optical modules and other detector components is synchronized at the level of sub-nanoseconds. The position of the module is measured by acoustic piezo detectors inside the module and external acoustic emitters installed on the bottom of the sea. The orientation of the module is obtained with an internal attitude and heading reference system chip. Detector calibration, i.e. timing, positioning and sea-water properties, is overviewed in this talk and discussed in detail in this conference. Results of the procedure applied to the first detector unit ready for installation in the deep sea will be shown.

  1. Comparison of two methods for high purity germanium detector efficiency calibration for charcoal canister radon measurement.

    PubMed

    Nikolic, J; Pantelic, G; Zivanovic, M; Rajacic, M; Todorovic, D

    2014-11-01

    The charcoal canister method of radon measurement according to US Environment Protection Agency protocol 520/5-87-005 is widely used for screening. This method is based on radon adsorption on coal and measurement of gamma radiation of radon daughters. For the purpose of gamma spectrometry, appropriate efficiency calibration of the measuring system must be performed. The most usual method of calibration is using standard canister, a sealed canister with the same matrix and geometry as the canisters used for measurements, but with the known activity of radon. In the absence of standard canister, a different method of efficiency calibration has to be implemented. This study presents the results of efficiency calibration using the EFFTRAN efficiency transfer software. Efficiency was calculated using a soil matrix cylindrical secondary reference material as a starting point. Calculated efficiency is then compared with the one obtained using standard canister and applied to a realistic measurement in order to evaluate the results of the efficiency transfer.

  2. A direct micropipette-based calibration method for atomic force microscope cantilevers

    PubMed Central

    Liu, Baoyu; Yu, Yan; Yao, Da-Kang; Shao, Jin-Yu

    2009-01-01

    In this report, we describe a direct method for calibrating atomic force microscope (AFM) cantilevers with the micropipette aspiration technique (MAT). A closely fitting polystyrene bead inside a micropipette is driven by precisely controlled hydrostatic pressures to apply known loads on the sharp tip of AFM cantilevers, thus providing a calibration at the most functionally relevant position. The new method is capable of calibrating cantilevers with spring constants ranging from 0.01 to hundreds of newtons per meter. Under appropriate loading conditions, this new method yields measurement accuracy and precision both within 10%, with higher performance for softer cantilevers. Furthermore, this method may greatly enhance the accuracy and precision of calibration for colloidal probes. PMID:19566228

  3. An efficient calibration method for freehand 3-D ultrasound imaging systems.

    PubMed

    Leotta, Daniel F

    2004-07-01

    A phantom has been developed to quickly calibrate a freehand 3-D ultrasound (US) imaging system. Calibration defines the spatial relationship between the US image plane and an external tracking device attached to the scanhead. The phantom consists of a planar array of strings and beads, and a set of out-of-plane strings that guide the user to proper scanhead orientation for imaging. When an US image plane is coincident with the plane defined by the strings, the calibration parameters are calculated by matching of homologous points in the image and phantom. The resulting precision and accuracy of the 3-D imaging system are similar to those achieved with a more complex calibration procedure. The 3-D reconstruction performance of the calibrated system is demonstrated with a magnetic tracking system, but the method could be applied to other tracking devices.

  4. SERI results from the PEP 1987 Summit Round Robin and a comparison of photovoltaic calibration methods

    SciTech Connect

    Emery, K.A.; Waddington, D.; Rummel, S.; Myers, D.R.; Stoffel, T.L.; Osterwald, C.R.

    1989-03-01

    The SERI results for the Photovoltaic Solar Energy Project (PEP) 1987 summit round robin are presented and the proposed International Electrotechnical Commission (IEC) technical committee TC82 global calibration method is evaluated in this report. The global calibration method has a +-4% uncertainty in the short-circuit current (I/sub sc/) including random and nonrandom error sources. This uncertainty can be reduced +-3% if the global normal method is used and the direct component of the total irradiance is measured with a primary absolute cavity radiometer. Nearly half of the uncertainty for various primary and secondary PV calibration methods is nonrandom. If a +-4% uncertainty in I/sub sc/ under standard reporting conditions is acceptable, then the global calibration method with spectral corrections is suitable for cells or modules. If a primary reference device with an uncertainty less than +-3% is required, then the global method is unsuitable. The SERI primary direct normal calibration method has a +-1% uncertainty in I/sub sc/ as verified by comparison with primary AMO cells. If a +-1% primary reference cell is used for secondary calibrations, then a +-2% uncertainty could be achieved for reference cells or modules. 37 refs., 29 figs., 42 tabs.

  5. METHOD OF OBTAINING AN ADDITIVE FOR LUBRICATING OILS,

    DTIC Science & Technology

    The method of obtaining an additive to lubricating oils , consisting of treating boron trifluoride with alkylphenol and alkylamine, is known. In this...case, (aminotrifluoralkyl)phenoxyborate is obtained which may be used as an antiwear additive for lubricating oils . The proposed method differs from

  6. A wire scanning based method for geometric calibration of high resolution CT system

    NASA Astrophysics Data System (ADS)

    Jiang, Ruijie; Li, Guang; Gu, Ning; Chen, Gong; Luo, Shouhua

    2015-03-01

    This paper is about geometric calibration of the high resolution CT (Computed Tomography) system. Geometric calibration refers to the estimation of a set of parameters that describe the geometry of the CT system. Such parameters are so important that a little error of them will degrade the reconstruction images seriously, so more accurate geometric parameters are needed in the higher-resolution CT systems. But conventional calibration methods are not accurate enough for the current high resolution CT system whose resolution can reach sub-micrometer or even tens of nanometers. In this paper, we propose a new calibration method which has higher accuracy and it is based on the optimization theory. The superiority of this method is that we build a new cost function which sets up a relationship between the geometrical parameters and the binary reconstruction image of a thin wire. When the geometrical parameters are accurate, the cost function reaches its maximum value. In the experiment, we scanned a thin wire as the calibration data and a thin bamboo stick as the validation data to verify the correctness of the proposed method. Comparing with the image reconstructed with the geometric parameters calculated by using the conventional calibration method, the image reconstructed with the parameters calculated by our method has less geometric artifacts, so it can verify that our method can get more accurate geometric calibration parameters. Although we calculated only one geometric parameter in this paper, the geometric artifacts are still eliminated significantly. And this method can be easily generalized to all the geometrical parameters calibration in fan-beam or cone-beam CT systems.

  7. Novel Method for Processing the Dynamic Calibration Signal of Pressure Sensor

    PubMed Central

    Wang, Zhongyu; Li, Qiang; Wang, Zhuoran; Yan, Hu

    2015-01-01

    Dynamic calibration is one of the important ways to acquire the dynamic performance parameters of a pressure sensor. This research focuses on the processing method for the output of calibrated pressure sensor, and mainly attempts to solve the problem of extracting the true information of step response under strong interference noise. A dynamic calibration system based on a shock tube is established to excite the time-domain response signal of a calibrated pressure sensor. A key processing on difference modeling is applied for the obtained signal, and several generating sequences are established. A fusion process for the generating sequences is then undertaken, and the true information of the step response of the calibrated pressure sensor can be obtained. Finally, by implementing the common QR decomposition method to deal with the true information, a dynamic model characterizing the dynamic performance of the calibrated pressure sensor is established. A typical pressure sensor was used to perform calibration tests and a frequency-domain experiment for the sensor was also conducted. Results show that the proposed method could effectively filter strong interference noise in the output of the sensor and the corresponding dynamic model could effectively characterize the dynamic performance of the pressure sensor. PMID:26197324

  8. Novel Method for Processing the Dynamic Calibration Signal of Pressure Sensor.

    PubMed

    Wang, Zhongyu; Li, Qiang; Wang, Zhuoran; Yan, Hu

    2015-07-21

    Dynamic calibration is one of the important ways to acquire the dynamic performance parameters of a pressure sensor. This research focuses on the processing method for the output of calibrated pressure sensor, and mainly attempts to solve the problem of extracting the true information of step response under strong interference noise. A dynamic calibration system based on a shock tube is established to excite the time-domain response signal of a calibrated pressure sensor. A key processing on difference modeling is applied for the obtained signal, and several generating sequences are established. A fusion process for the generating sequences is then undertaken, and the true information of the step response of the calibrated pressure sensor can be obtained. Finally, by implementing the common QR decomposition method to deal with the true information, a dynamic model characterizing the dynamic performance of the calibrated pressure sensor is established. A typical pressure sensor was used to perform calibration tests and a frequency-domain experiment for the sensor was also conducted. Results show that the proposed method could effectively filter strong interference noise in the output of the sensor and the corresponding dynamic model could effectively characterize the dynamic performance of the pressure sensor.

  9. Automatic Calibration Method for Driver’s Head Orientation in Natural Driving Environment

    PubMed Central

    Fu, Xianping; Guan, Xiao; Peli, Eli; Liu, Hongbo; Luo, Gang

    2013-01-01

    Gaze tracking is crucial for studying driver’s attention, detecting fatigue, and improving driver assistance systems, but it is difficult in natural driving environments due to nonuniform and highly variable illumination and large head movements. Traditional calibrations that require subjects to follow calibrators are very cumbersome to be implemented in daily driving situations. A new automatic calibration method, based on a single camera for determining the head orientation and which utilizes the side mirrors, the rear-view mirror, the instrument board, and different zones in the windshield as calibration points, is presented in this paper. Supported by a self-learning algorithm, the system tracks the head and categorizes the head pose in 12 gaze zones based on facial features. The particle filter is used to estimate the head pose to obtain an accurate gaze zone by updating the calibration parameters. Experimental results show that, after several hours of driving, the automatic calibration method without driver’s corporation can achieve the same accuracy as a manual calibration method. The mean error of estimated eye gazes was less than 5°in day and night driving. PMID:24639620

  10. Single Vector Calibration System for Multi-Axis Load Cells and Method for Calibrating a Multi-Axis Load Cell

    NASA Technical Reports Server (NTRS)

    Parker, Peter A. (Inventor)

    2003-01-01

    A single vector calibration system is provided which facilitates the calibration of multi-axis load cells, including wind tunnel force balances. The single vector system provides the capability to calibrate a multi-axis load cell using a single directional load, for example loading solely in the gravitational direction. The system manipulates the load cell in three-dimensional space, while keeping the uni-directional calibration load aligned. The use of a single vector calibration load reduces the set-up time for the multi-axis load combinations needed to generate a complete calibration mathematical model. The system also reduces load application inaccuracies caused by the conventional requirement to generate multiple force vectors. The simplicity of the system reduces calibration time and cost, while simultaneously increasing calibration accuracy.

  11. A method for in situ absolute DD yield calibration of neutron time-of-flight detectors on OMEGA using CR-39-based proton detectors

    DOE PAGES

    Waugh, C. J.; Rosenberg, M. J.; Zylstra, A. B.; ...

    2015-05-27

    Neutron time of flight (nTOF) detectors are used routinely to measure the absolute DD neutron yield at OMEGA. To check the DD yield calibration of these detectors, originally calibrated using indium activation systems, which in turn were cross-calibrated to NOVA nTOF detectors in the early 1990s, a direct in situ calibration method using CR-39 range filter proton detectors has been successfully developed. By measuring DD neutron and proton yields from a series of exploding pusher implosions at OMEGA, a yield calibration coefficient of 1.09 ± 0.02 (relative to the previous coefficient) was determined for the 3m nTOF detector. In addition,more » comparison of these and other shots indicates that significant reduction in charged particle flux anisotropies is achieved when bang time occurs significantly (on the order of 500 ps) after the trailing edge of the laser pulse. This is an important observation as the main source of the yield calibration error is due to particle anisotropies caused by field effects. The results indicate that the CR-39-nTOF in situ calibration method can serve as a valuable technique for calibrating and reducing the uncertainty in the DD absolute yield calibration of nTOF detector systems on OMEGA, the National Ignition Facility, and laser megajoule.« less

  12. A method for in situ absolute DD yield calibration of neutron time-of-flight detectors on OMEGA using CR-39-based proton detectors

    SciTech Connect

    Waugh, C. J.; Rosenberg, M. J.; Zylstra, A. B.; Frenje, J. A.; Seguin, F. H.; Petrasso, R. D.; Glebov, V. Yu.; Sangster, T. C.; Stoeckl, C.

    2015-05-27

    Neutron time of flight (nTOF) detectors are used routinely to measure the absolute DD neutron yield at OMEGA. To check the DD yield calibration of these detectors, originally calibrated using indium activation systems, which in turn were cross-calibrated to NOVA nTOF detectors in the early 1990s, a direct in situ calibration method using CR-39 range filter proton detectors has been successfully developed. By measuring DD neutron and proton yields from a series of exploding pusher implosions at OMEGA, a yield calibration coefficient of 1.09 ± 0.02 (relative to the previous coefficient) was determined for the 3m nTOF detector. In addition, comparison of these and other shots indicates that significant reduction in charged particle flux anisotropies is achieved when bang time occurs significantly (on the order of 500 ps) after the trailing edge of the laser pulse. This is an important observation as the main source of the yield calibration error is due to particle anisotropies caused by field effects. The results indicate that the CR-39-nTOF in situ calibration method can serve as a valuable technique for calibrating and reducing the uncertainty in the DD absolute yield calibration of nTOF detector systems on OMEGA, the National Ignition Facility, and laser megajoule.

  13. A method for in situ absolute DD yield calibration of neutron time-of-flight detectors on OMEGA using CR-39-based proton detectors

    SciTech Connect

    Waugh, C. J. Zylstra, A. B.; Frenje, J. A.; Séguin, F. H.; Petrasso, R. D.; Rosenberg, M. J.; Glebov, V. Yu.; Sangster, T. C.; Stoeckl, C.

    2015-05-15

    Neutron time of flight (nTOF) detectors are used routinely to measure the absolute DD neutron yield at OMEGA. To check the DD yield calibration of these detectors, originally calibrated using indium activation systems, which in turn were cross-calibrated to NOVA nTOF detectors in the early 1990s, a direct in situ calibration method using CR-39 range filter proton detectors has been successfully developed. By measuring DD neutron and proton yields from a series of exploding pusher implosions at OMEGA, a yield calibration coefficient of 1.09 ± 0.02 (relative to the previous coefficient) was determined for the 3m nTOF detector. In addition, comparison of these and other shots indicates that significant reduction in charged particle flux anisotropies is achieved when bang time occurs significantly (on the order of 500 ps) after the trailing edge of the laser pulse. This is an important observation as the main source of the yield calibration error is due to particle anisotropies caused by field effects. The results indicate that the CR-39-nTOF in situ calibration method can serve as a valuable technique for calibrating and reducing the uncertainty in the DD absolute yield calibration of nTOF detector systems on OMEGA, the National Ignition Facility, and laser megajoule.

  14. GMI Instrument Spin Balance Method, Optimization, Calibration, and Test

    NASA Technical Reports Server (NTRS)

    Ayari, Laoucet; Kubitschek, Michael; Ashton, Gunnar; Johnston, Steve; Debevec, Dave; Newell, David; Pellicciotti, Joseph

    2014-01-01

    The Global Microwave Imager (GMI) instrument must spin at a constant rate of 32 rpm continuously for the 3 year mission life. Therefore, GMI must be very precisely balanced about the spin axis and CG to maintain stable scan pointing and to minimize disturbances imparted to the spacecraft and attitude control on-orbit. The GMI instrument is part of the core Global Precipitation Measurement (GPM) spacecraft and is used to make calibrated radiometric measurements at multiple microwave frequencies and polarizations. The GPM mission is an international effort managed by the National Aeronautics and Space Administration (NASA) to improve climate, weather, and hydro-meteorological predictions through more accurate and frequent precipitation measurements. Ball Aerospace and Technologies Corporation (BATC) was selected by NASA Goddard Space Flight Center to design, build, and test the GMI instrument. The GMI design has to meet a challenging set of spin balance requirements and had to be brought into simultaneous static and dynamic spin balance after the entire instrument was already assembled and before environmental tests began. The focus of this contribution is on the analytical and test activities undertaken to meet the challenging spin balance requirements of the GMI instrument. The novel process of measuring the residual static and dynamic imbalances with a very high level of accuracy and precision is presented together with the prediction of the optimal balance masses and their locations.

  15. GMI Instrument Spin Balance Method, Optimization, Calibration and Test

    NASA Technical Reports Server (NTRS)

    Ayari, Laoucet; Kubitschek, Michael; Ashton, Gunnar; Johnston, Steve; Debevec, Dave; Newell, David; Pellicciotti, Joseph

    2014-01-01

    The Global Microwave Imager (GMI) instrument must spin at a constant rate of 32 rpm continuously for the 3-year mission life. Therefore, GMI must be very precisely balanced about the spin axis and center of gravity (CG) to maintain stable scan pointing and to minimize disturbances imparted to the spacecraft and attitude control on-orbit. The GMI instrument is part of the core Global Precipitation Measurement (GPM) spacecraft and is used to make calibrated radiometric measurements at multiple microwave frequencies and polarizations. The GPM mission is an international effort managed by the National Aeronautics and Space Administration (NASA) to improve climate, weather, and hydro-meteorological predictions through more accurate and frequent precipitation measurements. Ball Aerospace and Technologies Corporation (BATC) was selected by NASA Goddard Space Flight Center to design, build, and test the GMI instrument. The GMI design has to meet a challenging set of spin balance requirements and had to be brought into simultaneous static and dynamic spin balance after the entire instrument was already assembled and before environmental tests began. The focus of this contribution is on the analytical and test activities undertaken to meet the challenging spin balance requirements of the GMI instrument. The novel process of measuring the residual static and dynamic imbalances with a very high level of accuracy and precision is presented together with the prediction of the optimal balance masses and their locations.

  16. Method Specific Calibration Corrects for DNA Extraction Method Effects on Relative Telomere Length Measurements by Quantitative PCR.

    PubMed

    Seeker, Luise A; Holland, Rebecca; Underwood, Sarah; Fairlie, Jennifer; Psifidi, Androniki; Ilska, Joanna J; Bagnall, Ainsley; Whitelaw, Bruce; Coffey, Mike; Banos, Georgios; Nussey, Daniel H

    2016-01-01

    Telomere length (TL) is increasingly being used as a biomarker in epidemiological, biomedical and ecological studies. A wide range of DNA extraction techniques have been used in telomere experiments and recent quantitative PCR (qPCR) based studies suggest that the choice of DNA extraction method may influence average relative TL (RTL) measurements. Such extraction method effects may limit the use of historically collected DNA samples extracted with different methods. However, if extraction method effects are systematic an extraction method specific (MS) calibrator might be able to correct for them, because systematic effects would influence the calibrator sample in the same way as all other samples. In the present study we tested whether leukocyte RTL in blood samples from Holstein Friesian cattle and Soay sheep measured by qPCR was influenced by DNA extraction method and whether MS calibration could account for any observed differences. We compared two silica membrane-based DNA extraction kits and a salting out method. All extraction methods were optimized to yield enough high quality DNA for TL measurement. In both species we found that silica membrane-based DNA extraction methods produced shorter RTL measurements than the non-membrane-based method when calibrated against an identical calibrator. However, these differences were not statistically detectable when a MS calibrator was used to calculate RTL. This approach produced RTL measurements that were highly correlated across extraction methods (r > 0.76) and had coefficients of variation lower than 10% across plates of identical samples extracted by different methods. Our results are consistent with previous findings that popular membrane-based DNA extraction methods may lead to shorter RTL measurements than non-membrane-based methods. However, we also demonstrate that these differences can be accounted for by using an extraction method-specific calibrator, offering researchers a simple means of accounting for

  17. Method Specific Calibration Corrects for DNA Extraction Method Effects on Relative Telomere Length Measurements by Quantitative PCR

    PubMed Central

    Holland, Rebecca; Underwood, Sarah; Fairlie, Jennifer; Psifidi, Androniki; Ilska, Joanna J.; Bagnall, Ainsley; Whitelaw, Bruce; Coffey, Mike; Banos, Georgios; Nussey, Daniel H.

    2016-01-01

    Telomere length (TL) is increasingly being used as a biomarker in epidemiological, biomedical and ecological studies. A wide range of DNA extraction techniques have been used in telomere experiments and recent quantitative PCR (qPCR) based studies suggest that the choice of DNA extraction method may influence average relative TL (RTL) measurements. Such extraction method effects may limit the use of historically collected DNA samples extracted with different methods. However, if extraction method effects are systematic an extraction method specific (MS) calibrator might be able to correct for them, because systematic effects would influence the calibrator sample in the same way as all other samples. In the present study we tested whether leukocyte RTL in blood samples from Holstein Friesian cattle and Soay sheep measured by qPCR was influenced by DNA extraction method and whether MS calibration could account for any observed differences. We compared two silica membrane-based DNA extraction kits and a salting out method. All extraction methods were optimized to yield enough high quality DNA for TL measurement. In both species we found that silica membrane-based DNA extraction methods produced shorter RTL measurements than the non-membrane-based method when calibrated against an identical calibrator. However, these differences were not statistically detectable when a MS calibrator was used to calculate RTL. This approach produced RTL measurements that were highly correlated across extraction methods (r > 0.76) and had coefficients of variation lower than 10% across plates of identical samples extracted by different methods. Our results are consistent with previous findings that popular membrane-based DNA extraction methods may lead to shorter RTL measurements than non-membrane-based methods. However, we also demonstrate that these differences can be accounted for by using an extraction method-specific calibrator, offering researchers a simple means of accounting for

  18. Novel method and experimental validation of statistical calibration via Gaussianization in hot-wire anemometry

    NASA Astrophysics Data System (ADS)

    Gluzman, Igal; Cohen, Jacob; Oshman, Yaakov

    2016-11-01

    We introduce a statistical method based on Gaussianization to estimate the nonlinear calibration curve of a hot-wire probe, that relates the input flow velocity to the output (measured) voltage. The method uses as input a measured sequence of voltage samples, corresponding to different unknown flow velocities in the desired operational range, and only two measured voltages along with their known (calibrated) flow velocities. The novel method is validated against standard calibration methods using data acquired by hot-wire probes using wind-tunnel experiments. We demonstrate our new calibration technique by placing the hot-wire probe at certain region downstream of a cube-shaped body in a free stream of air flow. For testing our calibration method we rely on flow statistics that exist, among others, in a certain region of a turbulent wake formed downstream of the cube-shaped body. The specific properties are: first, the velocity signal in the wake should be as close to Gaussian as possible. Second, the signal should cover the desired velocity range that should be calibrated. The appropriate region to place our probe is determined via computation of the first four statistical moments of the measured signals in different regions of the wake.

  19. A new automated and precise calibration method for gamma level gauges with rod detector arrangement.

    PubMed

    Peyvandi, Reza Gholipour; Taheri, Ali; Olfateh, Ali; Islami, Seyyedeh Zahra

    2016-06-01

    Gamma-ray liquid level gauging is of particular importance in several industries. Industrial vessels, tanks, and reactors, which work at high temperatures and pressures, usually have thick metal walls up to 20cm. These factors make it impossible to know the exact level of liquid or fluid while the system is operating. For this reason, the calibration process of the gamma level gauges is difficult as it is impossible to gain access to the inside of the vessels, which is important during the calibration process. In this study, a new auto-calibration method was proposed for the aforementioned situations.

  20. Balance Calibration – A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance.

    SciTech Connect

    Mike Stears

    2010-07-01

    Paper Title: Balance Calibration – A method for assigning a direct-reading uncertainty to an electronic balance. Intended Audience: Those who calibrate or use electronic balances. Abstract: As a calibration facility, we provide on-site (at the customer’s location) calibrations of electronic balances for customers within our company. In our experience, most of our customers are not using their balance as a comparator, but simply putting an unknown quantity on the balance and reading the displayed mass value. Manufacturer’s specifications for balances typically include specifications such as readability, repeatability, linearity, and sensitivity temperature drift, but what does this all mean when the balance user simply reads the displayed mass value and accepts the reading as the true value? This paper discusses a method for assigning a direct-reading uncertainty to a balance based upon the observed calibration data and the environment where the balance is being used. The method requires input from the customer regarding the environment where the balance is used and encourages discussion with the customer regarding sources of uncertainty and possible means for improvement; the calibration process becomes an educational opportunity for the balance user as well as calibration personnel. This paper will cover the uncertainty analysis applied to the calibration weights used for the field calibration of balances; the uncertainty is calculated over the range of environmental conditions typically encountered in the field and the resulting range of air density. The temperature stability in the area of the balance is discussed with the customer and the temperature range over which the balance calibration is valid is decided upon; the decision is based upon the uncertainty needs of the customer and the desired rigor in monitoring by the customer. Once the environmental limitations are decided, the calibration is performed and the measurement data is entered into a

  1. A field calibration method to eliminate the error caused by relative tilt on roll angle measurement

    NASA Astrophysics Data System (ADS)

    Qi, Jingya; Wang, Zhao; Huang, Junhui; Yu, Bao; Gao, Jianmin

    2016-11-01

    The roll angle measurement method based on a heterodyne interferometer is an efficient technique for its high precision and environmental noise immunity. The optical layout bases on a polarization-assisted conversion of the roll angle into an optical phase shift, read by a beam passing through the objective plate actuated by the roll rotation. The measurement sensitivity or the gain coefficient G is calibrated before. However, a relative tilt between the laser and objective plate always exist due to the tilt of the laser and the roll of the guide in the field long rail measurement. The relative tilt affect the value of G, thus result in the roll angle measurement error. In this paper, a method for field calibration of G is presented to eliminate the measurement error above. The field calibration layout turns the roll angle into an optical path change (OPC) by a rotary table. Thus, the roll angle can be obtained from the OPC read by a two-frequency interferometer. Together with the phase shift, an accurate G in field measurement can be obtained and the measurement error can be corrected. The optical system of the field calibration method is set up and the experiment results are given. Contrasted with the Renishaw XL-80 for calibration, the proposed field calibration method can obtain the accurate G in the field rail roll angle measurement.

  2. A novel separation and calibration method for DVL and compass error in dead reckoning navigation systems

    NASA Astrophysics Data System (ADS)

    Zhang, Yanshun; Guo, Yajing; Yang, Tao; Li, Chunyu; Wang, Zhanqing

    2016-06-01

    The scale factor error δ C of the Doppler velocity log (DVL) and the heading angle error δ \\psi of a compass are so integrated in dead reckoning (DR) navigation systems that it is difficult to separate them. This paper aims to solve this problem by putting forward an online separation and calibration method for δ C and δ \\psi based on an ‘arc and linear’ trajectory. This method introduces the high-accuracy location information of a long base line (LBL) acoustic positioning system. At first, the relationship between the displacements on the ‘arc’ trajectory in directions of east and north, output by the LBL and DR systems, serves to judge the carrier direction and calibrate δ C . And then by compensating δ C , the displacement on the ‘linear’ trajectory is used to calibrate δ \\psi . Finally, a semi-physical simulation experiment is conducted to test and verify this calibration method to see how effective and accurate it is. Experimental results show that after calibration the residual error ratios of δ C and δ \\psi are 8.24% and 3.70% respectively. Therefore, online calibration of δ C and δ \\psi is realized effectively. What’s more, when the DR system is working alone in 400 s, this method reduces position error by up to 93.39%, from 18.91 m to 1.25 m.

  3. Method for Effective Calibration of Temperature Loggers with Automated Data Sampling and Evaluation

    NASA Astrophysics Data System (ADS)

    Ljungblad, S.; Josefson, L. E.; Holmsten, M.

    2011-12-01

    A highly automated calibration method for temperature loggers is presented. By using an automated procedure, a time- and cost-efficient calibration of temperature loggers is made possible. The method is directed at loggers that lack the function/property of direct reading from a display. This type of logger has to be connected to a computer for the setting-up of the measurement and again for collection of the measurement results. During the calibration, the loggers are offline. This method has been developed for temperature loggers from Gemini Data loggers, but the software and method could be modified to suit other types of loggers as well. Calibration is performed by comparison to a reference thermometer in liquid baths; and for loggers which have external sensors, only the sensor is normally placed in the bath. Loggers with internal sensors are protected from the liquid by placing them in an exterior plastic or metallic cover, and thereafter the entire loggers are placed in the bath. A digital thermometer measures the reference temperature of the bath and transmits it to a computer by way of Bluetooth. The developed calibration software, SPTempLogger, controls the logger software, and thus the communication protocol of the logger software does not need to be known. The previous method, with manual handling of the start and termination of every measuring sequence, evaluation of the resulting data and its corresponding uncertainty components, can be replaced by this automated method. Both the logger and reference measurement data are automatically downloaded once the logger has been connected to a computer after the calibration, and the calibration software started. The data are then evaluated automatically, and by statistical analysis of the confidence coefficient and standard deviation, the temperature plateaus that the calibration includes are identified. If a number of control parameters comply with the requirements, then the correction, resolution, and short

  4. A Radial Self-Calibrated (RASCAL) GRAPPA method using Weight Interpolation

    PubMed Central

    Codella, Noel C. F.; Spincemaille, Pascal; Prince, Martin; Wang, Yi

    2011-01-01

    A generalized autocalibrating partially parallel acquisition (GRAPPA) method for radial k-space sampling is presented that calculates GRAPPA weights without synthesized or acquired calibration data. Instead, GRAPPA weights are fit to the undersampled data as if it were the calibration data itself. Because the relative k-space shifts associated with these GRAPPA weights are varying for a radial trajectory, new GRAPPA weights can be resampled for arbitrary shifts through interpolation, which is then used to generate missing projections between the acquired projections. The method is demonstrated in phantoms and in abdominal and brain imaging. Image quality is similar to radial GRAPPA using fully sampled calibration data, and improved compared to a previously described self-calibrated radial GRAPPA technique. PMID:21834008

  5. Misalignment calibration of geomagnetic vector measurement system using parallelepiped frame rotation method

    NASA Astrophysics Data System (ADS)

    Pang, Hongfeng; Zhu, XueJun; Pan, Mengchun; Zhang, Qi; Wan, Chengbiao; Luo, Shitu; Chen, Dixiang; Chen, Jinfei; Li, Ji; Lv, Yunxiao

    2016-12-01

    Misalignment error is one key factor influencing the measurement accuracy of geomagnetic vector measurement system, which should be calibrated with the difficulties that sensors measure different physical information and coordinates are invisible. A new misalignment calibration method by rotating a parallelepiped frame is proposed. Simulation and experiment result show the effectiveness of calibration method. The experimental system mainly contains DM-050 three-axis fluxgate magnetometer, INS (inertia navigation system), aluminium parallelepiped frame, aluminium plane base. Misalignment angles are calculated by measured data of magnetometer and INS after rotating the aluminium parallelepiped frame on aluminium plane base. After calibration, RMS error of geomagnetic north, vertical and east are reduced from 349.441 nT, 392.530 nT and 562.316 nT to 40.130 nT, 91.586 nT and 141.989 nT respectively.

  6. A comprehensive uncertainty analysis and method of geometric calibration for a circular scanning airborne lidar

    NASA Astrophysics Data System (ADS)

    Gonsalves, Michael Oliver

    This dissertation describes an automated technique for ascertaining the values of the geometric calibration parameters of an airborne lidar. A least squares approach is employed that adjusts the point cloud to a single planar surface which could be either a narrow airport runway or a dynamic sea surface. Going beyond the customary three boresight angles, the proposed adjustment can determine up to eleven calibration parameters to a precision that renders a negligible contribution to the point cloud's positional uncertainty. Presently under development is the Coastal Zone Mapping and Imaging Lidar (CZMIL), which, unlike most contemporary systems that use oscillating mirrors to reflect the beam, will use a circular spinning prism to refract the laser in the desired direction. This departure from the traditional scanner presents the potential for internal geometric misalignments not previously experienced. Rather than relying on past calibration practices (like requiring data be acquired over a pitched-roof), a more robust method of calibration is established which does not depend on the presence of any cultural features. To develop this new method of calibration, the laser point positioning equation for this lidar was developed first. The system was then simulated in the MATLAB environment. Using these artificial datasets, the behavior of each geometric parameter iii was systematically manipulated, understood and calibrated, while an optimal flight strategy for the calibration acquisition was simultaneously developed. Finally, the total propagated uncertainty (TPU) of the point cloud was determined using a propagation of variances. Using this TPU module, the strength of the calibration solution was assessed. For example, four flight lines each of 20 seconds in duration contained sufficient information to determine the calibration parameters to such a degree of confidence that their contribution to the final point cloud uncertainty was only 0.012m in the horizontal

  7. A new and simple calibration-independent method for measuring the beam energy of a cyclotron.

    PubMed

    Gagnon, Katherine; Jensen, Mikael; Thisgaard, Helge; Publicover, Julia; Lapi, Suzanne; McQuarrie, Steve A; Ruth, Thomas J

    2011-01-01

    This work recommends a new and simple-to-perform method for measuring the beam energy of an accelerator. The proposed method requires the irradiation of two monitor foils interspaced by an energy degrader. The primary advantage of the proposed method, which makes this method unique from previous energy evaluation strategies that employ the use of monitor foils, is that this method is independent of the detector efficiency calibration. This method was evaluated by performing proton activation of (nat)Cu foils using both a cyclotron and a tandem Van de Graaff accelerator. The monitor foil activities were read using a dose calibrator set to an arbitrary calibration setting. Excellent agreement was noted between the nominal and measured proton energies.

  8. A new method for monitoring long term calibration of the SBUV and TOMS instruments

    NASA Technical Reports Server (NTRS)

    Ahmad, Z.; Seftor, C.; Wellemeyer, C.

    1994-01-01

    A new method has been developed to monitor the long-term calibration of the Solar Backscatter Ultraviolet (SBUV) and Total Ozone Mapping Spectrometer (TOMS) instruments. It is based on the fact that the radiance in one channel can be expressed as a linear sum of the radiances in neighboring channels. Using simulated radiances for the SBUV and TOMS instruments, various scenarios of changes in instrument calibration are investigated. Results from sample processing of SBUV data are also presented.

  9. Validation of the hypercapnic calibrated fMRI method using DOT-fMRI fusion imaging

    PubMed Central

    Yücel, Meryem A.; Evans, Karleyton C.; Selb, Juliette; Huppert, Theodore J.; Boas, David A.; Gagnon, Louis

    2014-01-01

    Calibrated functional Magnetic Resonance Imaging (fMRI) is a widely used method to investigate brain function in terms of physiological quantities such as the cerebral metabolic rate of oxygen (CMRO2). The first and one of the most common methods of fMRI calibration is hypercapnic calibration. This is achieved via simultaneous measures of blood-oxygenation-level dependent (BOLD) and the arterial spin labeling (ASL) signals during a functional task that evokes regional changes in CMRO2. A subsequent acquisition is then required during which the subject inhales carbon dioxide for short periods of time. A calibration constant, typically labeled M, is then estimated from the hypercapnic data and is subsequently used together with the BOLD-ASL recordings to compute evoked changes in CMRO2 during the functional task. The computation of M assumes a constant CMRO2 during the CO2 inhalation, an assumption that has been questioned since the origin of calibrated fMRI. In this study we used Diffuse Optical Tomography (DOT) together with BOLD and ASL – an alternative calibration method that does not require any gas manipulation and therefore no constant CMRO2 assumption - to cross-validate the estimation of M obtained from a traditional hypercapnic calibration. We found a high correlation between the M values (R=0.87, p<0.01) estimated using these two approaches. The findings serve to validate the hypercapnic fMRI calibration technique and suggest that the inter-subject variability routinely obtained for M is reproducible with an alternative method and might therefore reflect inter-subject physiological variability. PMID:25196509

  10. New methods of data calibration for high power-aperture lidar.

    PubMed

    Guan, Sai; Yang, Guotao; Chang, Qihai; Cheng, Xuewu; Yang, Yong; Gong, Shaohua; Wang, Jihong

    2013-03-25

    For high power-aperture lidar sounding of wide atmospheric dynamic ranges, as in middle-upper atmospheric probing, photomultiplier tubes' (PMT) pulse pile-up effects and signal-induced noise (SIN) complicates the extraction of information from lidar return signal, especially from metal layers' fluorescence signal. Pursuit for sophisticated description of metal layers' characteristics at far range (80~130km) with one PMT of high quantum efficiency (QE) and good SNR, contradicts the requirements for signals of wide linear dynamic range (i.e. from approximate 10(2) to 10(8) counts/s). In this article, Substantial improvements on experimental simulation of Lidar signals affected by PMT are reported to evaluate the PMTs' distortions in our High Power-Aperture Sodium LIDAR system. A new method for pile-up calibration is proposed by taking into account PMT and High Speed Data Acquisition Card as an Integrated Black-Box, as well as a new experimental method for identifying and removing SIN from the raw Lidar signals. Contradiction between the limited linear dynamic range of raw signal (55~80km) and requirements for wider acceptable linearity has been effectively solved, without complicating the current lidar system. Validity of these methods was demonstrated by applying calibrated data to retrieve atmospheric parameters (i.e. atmospheric density, temperature and sodium absolutely number density), in comparison with measurements of TIMED satellite and atmosphere model. Good agreements are obtained between results derived from calibrated signal and reference measurements where differences of atmosphere density, temperature are less than 5% in the stratosphere and less than 10K from 30km to mesosphere, respectively. Additionally, approximate 30% changes are shown in sodium concentration at its peak value. By means of the proposed methods to revert the true signal independent of detectors, authors approach a new balance between maintaining the linearity of adequate signal (20

  11. An aquaculture-based method for calibrated bivalve isotope paleothermometry

    NASA Astrophysics Data System (ADS)

    Wanamaker, Alan D.; Kreutz, Karl J.; Borns, Harold W.; Introne, Douglas S.; Feindel, Scott; Barber, Bruce J.

    2006-09-01

    To quantify species-specific relationships between bivalve carbonate isotope geochemistry (δ18Oc) and water conditions (temperature and salinity, related to water isotopic composition [δ18Ow]), an aquaculture-based methodology was developed and applied to Mytilus edulis (blue mussel). The four-by-three factorial design consisted of four circulating temperature baths (7, 11, 15, and 19°C) and three salinity ranges (23, 28, and 32 parts per thousand (ppt); monitored for δ18Ow weekly). In mid-July of 2003, 4800 juvenile mussels were collected in Salt Bay, Damariscotta, Maine, and were placed in each configuration. The size distribution of harvested mussels, based on 105 specimens, ranged from 10.9 mm to 29.5 mm with a mean size of 19.8 mm. The mussels were grown in controlled conditions for up to 8.5 months, and a paleotemperature relationship based on juvenile M. edulis from Maine was developed from animals harvested at months 4, 5, and 8.5. This relationship [T°C = 16.19 (±0.14) - 4.69 (±0.21) {δ18Oc VPBD - δ18Ow VSMOW} + 0.17 (±0.13) {δ18Oc VPBD - δ18Ow VSMOW}2; r2 = 0.99; N = 105; P < 0.0001] is nearly identical to the Kim and O'Neil (1997) abiogenic calcite equation over the entire temperature range (7-19°C), and it closely resembles the commonly used paleotemperature equations of Epstein et al. (1953) and Horibe and Oba (1972). Further, the comparison of the M. edulis paleotemperature equation with the Kim and O'Neil (1997) equilibrium-based equation indicates that M. edulis specimens used in this study precipitated their shell in isotopic equilibrium with ambient water within the experimental uncertainties of both studies. The aquaculture-based methodology described here allows similar species-specific isotope paleothermometer calibrations to be performed with other bivalve species and thus provides improved quantitative paleoenvironmental reconstructions.

  12. A visual servoing-based method for ProCam systems calibration.

    PubMed

    Berry, Francois; Aider, Omar Ait; Mosnier, Jeremie

    2013-10-01

    Projector-camera systems are currently used in a wide field of applications, such as 3D reconstruction and augmented reality, and can provide accurate measurements, depending on the configuration and calibration. Frequently, the calibration task is divided into two steps: camera calibration followed by projector calibration. The latter still poses certain problems that are not easy to solve, such as the difficulty in obtaining a set of 2D-3D points to compute the projection matrix between the projector and the world. Existing methods are either not sufficiently accurate or not flexible. We propose an easy and automatic method to calibrate such systems that consists in projecting a calibration pattern and superimposing it automatically on a known printed pattern. The projected pattern is provided by a virtual camera observing a virtual pattern in an OpenGL model. The projector displays what the virtual camera visualizes. Thus, the projected pattern can be controlled and superimposed on the printed one with the aid of visual servoing. Our experimental results compare favorably with those of other methods considering both usability and accuracy.

  13. Local Strategy Combined with a Wavelength Selection Method for Multivariate Calibration

    PubMed Central

    Chang, Haitao; Zhu, Lianqing; Lou, Xiaoping; Meng, Xiaochen; Guo, Yangkuan; Wang, Zhongyu

    2016-01-01

    One of the essential factors influencing the prediction accuracy of multivariate calibration models is the quality of the calibration data. A local regression strategy, together with a wavelength selection approach, is proposed to build the multivariate calibration models based on partial least squares regression. The local algorithm is applied to create a calibration set of spectra similar to the spectrum of an unknown sample; the synthetic degree of grey relation coefficient is used to evaluate the similarity. A wavelength selection method based on simple-to-use interactive self-modeling mixture analysis minimizes the influence of noisy variables, and the most informative variables of the most similar samples are selected to build the multivariate calibration model based on partial least squares regression. To validate the performance of the proposed method, ultraviolet-visible absorbance spectra of mixed solutions of food coloring analytes in a concentration range of 20–200 µg/mL is measured. Experimental results show that the proposed method can not only enhance the prediction accuracy of the calibration model, but also greatly reduce its complexity. PMID:27271636

  14. Gafchromic EBT2 film dosimetry in reflection mode with a novel plan-based calibration method

    SciTech Connect

    Mendez, I.; Hartman, V.; Hudej, R.; Strojnik, A.; Casar, B.

    2013-01-15

    Purpose:A dosimetric system formed by Gafchromic EBT2 radiochromic film and Epson Expression 10000XL flatbed scanner was commissioned for dosimetry. In this paper, several open questions concerning the commissioning of radiochromic films for dosimetry were addressed: (a) is it possible to employ this dosimetric system in reflection mode; (b) if so, can the methods used in transmission mode also be used in reflection mode; (c) is it possible to obtain accurate absolute dose measurements with Gafchromic EBT2 films; (d) which calibration method should be followed; (e) which calibration models should be used; and (f) does three-color channel dosimetry offer a significant improvement over single channel dosimetry. The purpose of this paper is to help clarify these questions. Methods: In this study, films were scanned in reflection mode, the effect of surrounding film was evaluated and the feasibility of EBT2 film dosimetry in reflection mode was studied. EBT2's response homogeneity has been reported to lead to excessive dose uncertainties. To overcome this problem, a new plan-based calibration method was implemented. Plan-based calibration can use every pixel and each of the three color channels of the scanned film to obtain the parameters of the calibration model. A model selection analysis was conducted to select lateral correction and sensitometric curve models. The commonly used calibration with fragments was compared with red-channel plan-based calibration and with three-channel plan-based calibration. Results: No effect of surrounding film was found in this study. The film response inhomogeneity in EBT2 films was found to be important not only due to differences in the fog but also due to differences in sensitivity. The best results for lateral corrections were obtained using absolute corrections independent of the dose. With respect to the sensitometric curves, an empirical polynomial fit of order 4 was found to obtain results equivalent to a gamma

  15. High-precision self-adaptive phase-calibration method for wavelength-tuning interferometry

    NASA Astrophysics Data System (ADS)

    Zhu, Xueliang; Zhao, Huiying; Dong, Longchao; Wang, Hongjun; Liu, Bingcai; Yuan, Daocheng; Tian, Ailing; Wang, Fangjie; Zhang, Chupeng; Ban, Xinxing

    2017-03-01

    We introduce a high-precision self-adaptive phase-calibration method for performing wavelength-tuning interferometry. Our method is insensitive to the nonlinearity of the phase shifter, even under random control. Intensity errors derived from laser voltage changes can be restrained by adopting this approach. Furthermore, this method can effectively overcome the influences from the background and modulation intensities in the interferogram, regardless of the phase structure. Numerical simulations and experiments are implemented to verify the validity of this high-precision calibration method.

  16. Data Sparseness and Online Pretest Item Calibration/Scaling Methods in CAT. ACT Research Report Series.

    ERIC Educational Resources Information Center

    Ban, Jae-Chun; Hanson, Bradley A.; Yi, Qing; Harris, Deborah J.

    The purpose of this study was to compare and evaluate three online pretest item calibration/scaling methods in terms of item parameter recovery when the item responses to the pretest items in the pool would be sparse. The three methods considered were the marginal maximum likelihood estimate with one EM cycle (OEM) method, the marginal maximum…

  17. Data Sparseness and On-Line Pretest Item Calibration-Scaling Methods in CAT.

    ERIC Educational Resources Information Center

    Ban, Jae-Chun; Hanson, Bradley A.; Yi, Qing; Harris, Deborah J.

    2002-01-01

    Compared three online pretest calibration scaling methods through simulation: (1) marginal maximum likelihood with one expectation maximization (EM) cycle (OEM) method; (2) marginal maximum likelihood with multiple EM cycles (MEM); and (3) M. Stocking's method B. MEM produced the smallest average total error in parameter estimation; OEM yielded…

  18. Calibration of the straightness and orthogonality error of a laser feedback high-precision stage using self-calibration methods

    NASA Astrophysics Data System (ADS)

    Kim, Dongmin; Kim, Kihyun; Park, Sang Hyun; Jang, Sangdon

    2014-12-01

    An ultra high-precision 3-DOF air-bearing stage is developed and calibrated in this study. The stage was developed for the transportation of a glass or wafer with x and y following errors in the nanometer regime. To apply the proposed stage to display or semiconductor fabrication equipment, x and y straightness errors should be at the sub-micron level and the x-y orthogonality error should be in the region of several arcseconds with strokes of several hundreds of mm. Our system was designed to move a 400 mm stroke on the x axis and a 700 mm stroke on the y axis. To do this, 1000 mm and 550 mm bar-type mirrors were adopted for real time Δx and Δy laser measurements and feedback control. In this system, with the laser wavelength variation and instability being kept to a minimum through environmental control, the straightness and orthogonality become purely dependent upon the surface shape of the bar mirrors. Compensation for the distortion of the bar mirrors is accomplished using a self-calibration method. The successful application of the method nearly eliminated the straightness and orthogonality errors of the stage, allowing their specifications to be fully satisfied. As a result, the straightness and orthogonality errors of the stage were successfully decreased from 4.4 μm to 0.8 μm and from 0.04° to 2.48 arcsec, respectively.

  19. Soil moisture sensing with cosmic ray neutron probes: Do forests and other low-count locations call for additional calibration?

    NASA Astrophysics Data System (ADS)

    Heidbüchel, Ingo; Güntner, Andreas; Blume, Theresa

    2015-04-01

    Cosmic ray neutron sensors are bridging the scale gap between point measurements and remotely sensed data of soil moisture. We present a best practice for the calibration of cosmic ray neutron sensors in areas with low count rates (<1000 counts/hour). These areas can include low altitude and high latitude regions as well as forested areas where the sensor is situated below the canopy. After calibrating a sensor ten times within a single year it became clear that individual calibrations at different points in time yield a soil moisture-neutron count relationship that cannot be described by the Desilets et al. (2010) calibration function. Neither can it be explained by the dynamics in canopy biomass (we analyzed the influence of growing and falling leaves and water redistribution within the vegetation surrounding the instrument). Our empirically determined function appears to be flatter, i.e. a decrease in soil moisture causes a smaller increase in neutron counts than would be expected from the Desilets calibration function. Therefore we cannot assume that the calibration function has a constant shape irrespective of the measurement location. In order to yield consistent soil moisture measurements we thus recommend a two-point calibration with one of the calibrations being performed during high soil moisture conditions and a second one during low soil moisture conditions.

  20. An efficient calibration method for SQUID measurement system using three orthogonal Helmholtz coils

    NASA Astrophysics Data System (ADS)

    Hua, Li; Shu-Lin, Zhang; Chao-Xiang, Zhang; Xiang-Yan, Kong; Xiao-Ming, Xie

    2016-06-01

    For a practical superconducting quantum interference device (SQUID) based measurement system, the Tesla/volt coefficient must be accurately calibrated. In this paper, we propose a highly efficient method of calibrating a SQUID magnetometer system using three orthogonal Helmholtz coils. The Tesla/volt coefficient is regarded as the magnitude of a vector pointing to the normal direction of the pickup coil. By applying magnetic fields through a three-dimensional Helmholtz coil, the Tesla/volt coefficient can be directly calculated from magnetometer responses to the three orthogonally applied magnetic fields. Calibration with alternating current (AC) field is normally used for better signal-to-noise ratio in noisy urban environments and the results are compared with the direct current (DC) calibration to avoid possible effects due to eddy current. In our experiment, a calibration relative error of about 6.89 × 10-4 is obtained, and the error is mainly caused by the non-orthogonality of three axes of the Helmholtz coils. The method does not need precise alignment of the magnetometer inside the Helmholtz coil. It can be used for the multichannel magnetometer system calibration effectively and accurately. Project supported by the “Strategic Priority Research Program (B)” of the Chinese Academy of Sciences (Grant No. XDB04020200) and the Shanghai Municipal Science and Technology Commission Project, China (Grant No. 15DZ1940902).

  1. Reconciling Empirical Carbonate Clumped Isotope Calibrations: A Comparison of Calcite Precipitation and Acid Digestion Methods

    NASA Astrophysics Data System (ADS)

    Kelson, J.; Huntington, K. W.; Schauer, A. J.; Saenger, C.; Lechler, A. R.

    2015-12-01

    An accurate empirical calibration is necessary to confidently apply the carbonate clumped isotope (Δ47) thermometer. Previous synthetic carbonate calibrations disagree in temperature sensitivity, with one group of calibrations displaying a shallow Δ47-temperature slope (e.g., Dennis & Schrag, GCA, 2010), and the other a steep slope (e.g., Zaarur et al., EPSL, 2013). These calibrations differ in both the method of mineral precipitation and the temperature of the phosphoric acid used to digest carbonates for analysis, making it difficult to isolate the cause of the discrepancy. Here, we precipitate synthetic carbonates at temperatures of 6-80ºC using 4 different precipitation methods, and analyze the samples using both 90 and 25°C acid digestion. Precipitation experiments varied the use of salts (NaHCO3 and CaCl2) vs. dissolved CaCO3 as a starting solution, the use of carbonic anhydrase to promote isotopic equilibrium among dissolved inorganic carbon species in solution, and the method by which CO2 degasses to force carbonate precipitation. Carbonates precipitated by using salts and allowing CO2 to passively degas produce a shallow calibration slope that we hypothesize to approach isotopic equilibrium. Precipitation methods that bubble CO2 into solution then degas that CO2 (either passively or actively by bubbling N2) produce carbonates with consistently lower Δ47 and higher δ18O values for a given growth temperature. We infer that these carbonates grew in disequilibrium during rapid CO2 degassing. Varying acid digestion temperature does not change the results; acid fractionation factor is not correlated with grain size, Δ47, or d47 values. No precipitation method produces a steep calibration slope. Our large sample set of >60 carbonates lend confidence to a shallow slope calibration, and inform interpretations of Δ47 and δ18O values of natural carbonates that grow under conditions of isotopic disequilibrium.

  2. Crystal timing offset calibration method for time of flight PET scanners

    NASA Astrophysics Data System (ADS)

    Ye, Jinghan; Song, Xiyun

    2016-03-01

    In time-of-flight (TOF) positron emission tomography (PET), precise calibration of the timing offset of each crystal of a PET scanner is essential. Conventionally this calibration requires a specially designed tool just for this purpose. In this study a method that uses a planar source to measure the crystal timing offsets (CTO) is developed. The method uses list mode acquisitions of a planar source placed at multiple orientations inside the PET scanner field-of-view (FOV). The placement of the planar source in each acquisition is automatically figured out from the measured data, so that a fixture for exactly placing the source is not required. The expected coincidence time difference for each detected list mode event can be found from the planar source placement and the detector geometry. A deviation of the measured time difference from the expected one is due to CTO of the two crystals. The least squared solution of the CTO is found iteratively using the list mode events. The effectiveness of the crystal timing calibration method is evidenced using phantom images generated by placing back each list mode event into the image space with the timing offset applied to each event. The zigzagged outlines of the phantoms in the images become smooth after the crystal timing calibration is applied. In conclusion, a crystal timing calibration method is developed. The method uses multiple list mode acquisitions of a planar source to find the least squared solution of crystal timing offsets.

  3. Error analyses and calibration methods with accelerometers for optical angle encoders in rotational inertial navigation systems.

    PubMed

    Liu, Fang; Wang, Wei; Wang, Lei; Feng, Peide

    2013-11-10

    By rotating a strapdown inertial navigation system (INS) over one or more axes, a number of error sources originating from the employed sensors cancel out during the integration process. Rotary angle accuracy has an effect on the performance of rotational INS (RINS). The application of existing calibration methods based on gyroscope measurements is restricted by the structure of the inertial measurement unit (IMU) and scale factor stability of the gyroscope. The multireadhead method has problems in miniaturization and cost. Hence, optical angle encoder calibration methods using accelerometers are proposed, on the basis of navigation error and accuracy requirement analyses for a single-axis RINS. The test results show that the accuracy of calibration methods proposed is higher than 4 arcsec (1σ).

  4. Evaluation of Linking Methods for Multidimensional IRT Calibrations

    ERIC Educational Resources Information Center

    Min, Kyung-Seok

    2007-01-01

    Most researchers agree that psychological/educational tests are sensitive to multiple traits, implying the need for a multidimensional item response theory (MIRT). One limitation of applying a MIRT in practice is the difficulty in establishing equivalent scales of multiple traits. In this study, a new MIRT linking method was proposed and evaluated…

  5. Additives

    NASA Technical Reports Server (NTRS)

    Smalheer, C. V.

    1973-01-01

    The chemistry of lubricant additives is discussed to show what the additives are chemically and what functions they perform in the lubrication of various kinds of equipment. Current theories regarding the mode of action of lubricant additives are presented. The additive groups discussed include the following: (1) detergents and dispersants, (2) corrosion inhibitors, (3) antioxidants, (4) viscosity index improvers, (5) pour point depressants, and (6) antifouling agents.

  6. A GPS-Based Pitot-Static Calibration Method Using Global Output-Error Optimization

    NASA Technical Reports Server (NTRS)

    Foster, John V.; Cunningham, Kevin

    2010-01-01

    Pressure-based airspeed and altitude measurements for aircraft typically require calibration of the installed system to account for pressure sensing errors such as those due to local flow field effects. In some cases, calibration is used to meet requirements such as those specified in Federal Aviation Regulation Part 25. Several methods are used for in-flight pitot-static calibration including tower fly-by, pacer aircraft, and trailing cone methods. In the 1990 s, the introduction of satellite-based positioning systems to the civilian market enabled new inflight calibration methods based on accurate ground speed measurements provided by Global Positioning Systems (GPS). Use of GPS for airspeed calibration has many advantages such as accuracy, ease of portability (e.g. hand-held) and the flexibility of operating in airspace without the limitations of test range boundaries or ground telemetry support. The current research was motivated by the need for a rapid and statistically accurate method for in-flight calibration of pitot-static systems for remotely piloted, dynamically-scaled research aircraft. Current calibration methods were deemed not practical for this application because of confined test range size and limited flight time available for each sortie. A method was developed that uses high data rate measurements of static and total pressure, and GPSbased ground speed measurements to compute the pressure errors over a range of airspeed. The novel application of this approach is the use of system identification methods that rapidly compute optimal pressure error models with defined confidence intervals in nearreal time. This method has been demonstrated in flight tests and has shown 2- bounds of approximately 0.2 kts with an order of magnitude reduction in test time over other methods. As part of this experiment, a unique database of wind measurements was acquired concurrently with the flight experiments, for the purpose of experimental validation of the

  7. A new rapid method of solar simulator calibration

    NASA Technical Reports Server (NTRS)

    Ross, B.

    1976-01-01

    A quick method for checking solar simulator spectra content is presented. The method is based upon a solar cell of extended spectral sensitivity and known absolute response, and a dichroic mirror with the reflection transmission transition close to the peak wavelength of the Thekaekara AMO distribution. It compromises the need for spectral discrimination with the ability to integrate wide spectral regions of the distribution which was considered important due to the spiky nature of the high pressure xenon lamp in common use. The results are expressed in terms of a single number, the blue/red ratio, which, combined with the total (unfiltered) output, provides a simple adequate characterization. Measurements were conducted at eleven major facilities across the country and a total of eighteen simulators were measured including five pulsed units.

  8. A new time calibration method for switched-capacitor-array-based waveform samplers

    SciTech Connect

    Kim, H.; Chen, C. -T.; Eclov, N.; Ronzhin, A.; Murat, P.; Ramberg, E.; Los, S.; Moses, W.; Choong, W. -S.; Kao, C. -M.

    2014-08-24

    Here we have developed a new time calibration method for the DRS4 waveform sampler that enables us to precisely measure the non-uniform sampling interval inherent in the switched-capacitor cells of the DRS4. The method uses the proportionality between the differential amplitude and sampling interval of adjacent switched-capacitor cells responding to a sawtooth-shape pulse. In the experiment, a sawtooth-shape pulse with a 40 ns period generated by a Tektronix AWG7102 is fed to a DRS4 evaluation board for calibrating the sampling intervals of all 1024 cells individually. The electronic time resolution of the DRS4 evaluation board with the new time calibration is measured to be ~2.4 ps RMS by using two simultaneous Gaussian pulses with 2.35 ns full-width at half-maximum and applying a Gaussian fit. The time resolution dependencies on the time difference with the new time calibration are measured and compared to results obtained by another method. Ultimately, the new method could be applicable for other switched-capacitor-array technology-based waveform samplers for precise time calibration.

  9. A method for the temperature calibration of an infrared camera using water as a radiative source

    SciTech Connect

    Bower, S. M.; Kou, J.; Saylor, J. R.

    2009-09-15

    Presented here is an effective low-cost method for the temperature calibration of infrared cameras, for applications in the 0-100 deg. C range. The calibration of image gray level intensity to temperature is achieved by imaging an upwelling flow of water, the temperature of which is measured with a thermistor probe. The upwelling flow is created by a diffuser located below the water surface of a constant temperature water bath. The thermistor probe is kept immediately below the surface, and the distance from the diffuser outlet to the surface is adjusted so that the deformation of the water surface on account of the flow is small, yet the difference between the surface temperature seen by the camera and the bulk temperature measured by the thermistor is also small. The benefit of this method compared to typical calibration procedures is that, without sacrificing the quality of the calibration, relatively expensive commercial blackbodies are replaced by water as the radiative source ({epsilon}{approx_equal}0.98 for the wavelengths considered here). A heat transfer analysis is provided, which improves the accuracy of the calibration method and also provides the user with guidance to further increases in accuracy of the method.

  10. [Outlier sample discriminating methods for building calibration model in melons quality detecting using NIR spectra].

    PubMed

    Tian, Hai-Qing; Wang, Chun-Guang; Zhang, Hai-Jun; Yu, Zhi-Hong; Li, Jian-Kang

    2012-11-01

    Outlier samples strongly influence the precision of the calibration model in soluble solids content measurement of melons using NIR Spectra. According to the possible sources of outlier samples, three methods (predicted concentration residual test; Chauvenet test; leverage and studentized residual test) were used to discriminate these outliers respectively. Nine suspicious outliers were detected from calibration set which including 85 fruit samples. Considering the 9 suspicious outlier samples maybe contain some no-outlier samples, they were reclaimed to the model one by one to see whether they influence the model and prediction precision or not. In this way, 5 samples which were helpful to the model joined in calibration set again, and a new model was developed with the correlation coefficient (r) 0. 889 and root mean square errors for calibration (RMSEC) 0.6010 Brix. For 35 unknown samples, the root mean square errors prediction (RMSEP) was 0.854 degrees Brix. The performance of this model was more better than that developed with non outlier was eliminated from calibration set (r = 0.797, RMSEC= 0.849 degrees Brix, RMSEP = 1.19 degrees Brix), and more representative and stable with all 9 samples were eliminated from calibration set (r = 0.892, RMSEC = 0.605 degrees Brix, RMSEP = 0.862 degrees).

  11. Additive manufacturing method for SRF components of various geometries

    SciTech Connect

    Rimmer, Robert; Frigola, Pedro E; Murokh, Alex Y

    2015-05-05

    An additive manufacturing method for forming nearly monolithic SRF niobium cavities and end group components of arbitrary shape with features such as optimized wall thickness and integral stiffeners, greatly reducing the cost and technical variability of conventional cavity construction. The additive manufacturing method for forming an SRF cavity, includes atomizing niobium to form a niobium powder, feeding the niobium powder into an electron beam melter under a vacuum, melting the niobium powder under a vacuum in the electron beam melter to form an SRF cavity; and polishing the inside surface of the SRF cavity.

  12. Characterisation methods for the hyperspectral sensor HySpex at DLR's calibration home base

    NASA Astrophysics Data System (ADS)

    Baumgartner, Andreas; Gege, Peter; Köhler, Claas; Lenhard, Karim; Schwarzmaier, Thomas

    2012-09-01

    The German Aerospace Center's (DLR) Remote Sensing Technology Institute (IMF) operates a laboratory for the characterisation of imaging spectrometers. Originally designed as Calibration Home Base (CHB) for the imaging spectrometer APEX, the laboratory can be used to characterise nearly every airborne hyperspectral system. Characterisation methods will be demonstrated exemplarily with HySpex, an airborne imaging spectrometer system from Norsk Elektro Optikks A/S (NEO). Consisting of two separate devices (VNIR-1600 and SWIR-320me) the setup covers the spectral range from 400 nm to 2500 nm. Both airborne sensors have been characterised at NEO. This includes measurement of spectral and spatial resolution and misregistration, polarisation sensitivity, signal to noise ratios and the radiometric response. The same parameters have been examined at the CHB and were used to validate the NEO measurements. Additionally, the line spread functions (LSF) in across and along track direction and the spectral response functions (SRF) for certain detector pixels were measured. The high degree of lab automation allows the determination of the SRFs and LSFs for a large amount of sampling points. Despite this, the measurement of these functions for every detector element would be too time-consuming as typical detectors have 105 elements. But with enough sampling points it is possible to interpolate the attributes of the remaining pixels. The knowledge of these properties for every detector element allows the quantification of spectral and spatial misregistration (smile and keystone) and a better calibration of airborne data. Further laboratory measurements are used to validate the models for the spectral and spatial properties of the imaging spectrometers. Compared to the future German spaceborne hyperspectral Imager EnMAP, the HySpex sensors have the same or higher spectral and spatial resolution. Therefore, airborne data will be used to prepare for and validate the spaceborne system

  13. Standardization of Laser Methods and Techniques for Vibration Measurements and Calibrations

    SciTech Connect

    Martens, Hans-Juergen von

    2010-05-28

    The realization and dissemination of the SI units of motion quantities (vibration and shock) have been based on laser interferometer methods specified in international documentary standards. New and refined laser methods and techniques developed by national metrology institutes and by leading manufacturers in the past two decades have been swiftly specified as standard methods for inclusion into in the series ISO 16063 of international documentary standards. A survey of ISO Standards for the calibration of vibration and shock transducers demonstrates the extended ranges and improved accuracy (measurement uncertainty) of laser methods and techniques for vibration and shock measurements and calibrations. The first standard for the calibration of laser vibrometers by laser interferometry or by a reference accelerometer calibrated by laser interferometry (ISO 16063-41) is on the stage of a Draft International Standard (DIS) and may be issued by the end of 2010. The standard methods with refined techniques proved to achieve wider measurement ranges and smaller measurement uncertainties than that specified in the ISO Standards. The applicability of different standardized interferometer methods to vibrations at high frequencies was recently demonstrated up to 347 kHz (acceleration amplitudes up to 350 km/s{sup 2}). The relative deviations between the amplitude measurement results of the different interferometer methods that were applied simultaneously, differed by less than 1% in all cases.

  14. Method for pan-tilt camera calibration using single control point.

    PubMed

    Li, Yunting; Zhang, Jun; Hu, Wenwen; Tian, Jinwen

    2015-01-01

    The pan-tilt (PT) camera is widely used in video surveillance systems due to its rotatable property and low cost. The rough output of a PT camera may not satisfy the demand of practical applications; hence an accurate calibration method of a PT camera is desired. However, high-precision camera calibration methods usually require sufficient control points not guaranteed in some practical cases of a PT camera. In this paper, we present a novel method to online calibrate the rotation angles of a PT camera by using only one control point. This is achieved by assuming that the intrinsic parameters and position of the camera are known in advance. More specifically, we first build a nonlinear PT camera model with respect to two parameters Pan and Tilt. We then convert the nonlinear model into a linear model according to sine and cosine of Tilt, where each element in the augmented coefficient matrix is a function of the single variable Pan. A closed-form solution of Pan and Tilt can then be derived by solving a quadratic equation of tangent of Pan. Our method is noniterative and does not need features matching; thus its time efficiency is better. We evaluate our calibration method on various synthetic and real data. The quantitative results demonstrate that the proposed method outperforms other state-of-the-art methods if the intrinsic parameters and position of the camera are known in advance.

  15. Evaluation of calibration methods for visible-spectrum division-of-focal-plane polarimeters

    NASA Astrophysics Data System (ADS)

    Powell, S. B.; Gruev, Viktor

    2013-09-01

    Polarization imaging sensors using the division-of-focal-plane paradigm have recently emerged on the market. These sensors, due to their compact design, are ideal for field work. One of the major drawbacks in these sensors is the spatial variation of the optical response of individual pixels across the imaging array. These spatial variations are due to variations in the nanowires of the pixelated polarization filters. In this paper, we describe and compare two methods for calibrating a division of focal plane sensors. We present theoretical and experimental data for these calibration methods.

  16. A geometric calibration method for inverse geometry computed tomography using P-matrices

    NASA Astrophysics Data System (ADS)

    Slagowski, Jordan M.; Dunkerley, David A. P.; Hatt, Charles R.; Speidel, Michael A.

    2016-03-01

    Accurate and artifact free reconstruction of tomographic images requires precise knowledge of the imaging system geometry. This work proposes a novel projection matrix (P-matrix) based calibration method to enable C-arm inverse geometry CT (IGCT). The method is evaluated for scanning-beam digital x-ray (SBDX), a C-arm mounted inverse geometry fluoroscopic technology. A helical configuration of fiducials is imaged at each gantry angle in a rotational acquisition. For each gantry angle, digital tomosynthesis is performed at multiple planes and a composite image analogous to a cone-beam projection is generated from the plane stack. The geometry of the C-arm, source array, and detector array is determined at each angle by constructing a parameterized 3D-to-2D projection matrix that minimizes the sum-of-squared deviations between measured and projected fiducial coordinates. Simulations were used to evaluate calibration performance with translations and rotations of the source and detector. In a geometry with 1 mm translation of the central ray relative to the axis-of-rotation and 1 degree yaw of the detector and source arrays, the maximum error in the recovered translational parameters was 0.4 mm and maximum error in the rotation parameter was 0.02 degrees. The relative rootmean- square error in a reconstruction of a numerical thorax phantom was 0.4% using the calibration method, versus 7.7% without calibration. Changes in source-detector-distance were the most challenging to estimate. Reconstruction of experimental SBDX data using the proposed method eliminated double contour artifacts present in a non-calibrated reconstruction. The proposed IGCT geometric calibration method reduces image artifacts when uncertainties exist in system geometry.

  17. A geometric calibration method for inverse geometry computed tomography using P-matrices.

    PubMed

    Slagowski, Jordan M; Dunkerley, David A P; Hatt, Charles R; Speidel, Michael A

    2016-02-27

    Accurate and artifact free reconstruction of tomographic images requires precise knowledge of the imaging system geometry. This work proposes a novel projection matrix (P-matrix) based calibration method to enable C-arm inverse geometry CT (IGCT). The method is evaluated for scanning-beam digital x-ray (SBDX), a C-arm mounted inverse geometry fluoroscopic technology. A helical configuration of fiducials is imaged at each gantry angle in a rotational acquisition. For each gantry angle, digital tomosynthesis is performed at multiple planes and a composite image analogous to a cone-beam projection is generated from the plane stack. The geometry of the C-arm, source array, and detector array is determined at each angle by constructing a parameterized 3D-to-2D projection matrix that minimizes the sum-of-squared deviations between measured and projected fiducial coordinates. Simulations were used to evaluate calibration performance with translations and rotations of the source and detector. In a geometry with 1 mm translation of the central ray relative to the axis-of-rotation and 1 degree yaw of the detector and source arrays, the maximum error in the recovered translational parameters was 0.4 mm and maximum error in the rotation parameter was 0.02 degrees. The relative root-mean-square error in a reconstruction of a numerical thorax phantom was 0.4% using the calibration method, versus 7.7% without calibration. Changes in source-detector-distance were the most challenging to estimate. Reconstruction of experimental SBDX data using the proposed method eliminated double contour artifacts present in a non-calibrated reconstruction. The proposed IGCT geometric calibration method reduces image artifacts when uncertainties exist in system geometry.

  18. A geometric calibration method for inverse geometry computed tomography using P-matrices

    PubMed Central

    Slagowski, Jordan M.; Dunkerley, David A. P.; Hatt, Charles R.; Speidel, Michael A.

    2016-01-01

    Accurate and artifact free reconstruction of tomographic images requires precise knowledge of the imaging system geometry. This work proposes a novel projection matrix (P-matrix) based calibration method to enable C-arm inverse geometry CT (IGCT). The method is evaluated for scanning-beam digital x-ray (SBDX), a C-arm mounted inverse geometry fluoroscopic technology. A helical configuration of fiducials is imaged at each gantry angle in a rotational acquisition. For each gantry angle, digital tomosynthesis is performed at multiple planes and a composite image analogous to a cone-beam projection is generated from the plane stack. The geometry of the C-arm, source array, and detector array is determined at each angle by constructing a parameterized 3D-to-2D projection matrix that minimizes the sum-of-squared deviations between measured and projected fiducial coordinates. Simulations were used to evaluate calibration performance with translations and rotations of the source and detector. In a geometry with 1 mm translation of the central ray relative to the axis-of-rotation and 1 degree yaw of the detector and source arrays, the maximum error in the recovered translational parameters was 0.4 mm and maximum error in the rotation parameter was 0.02 degrees. The relative root-mean-square error in a reconstruction of a numerical thorax phantom was 0.4% using the calibration method, versus 7.7% without calibration. Changes in source-detector-distance were the most challenging to estimate. Reconstruction of experimental SBDX data using the proposed method eliminated double contour artifacts present in a non-calibrated reconstruction. The proposed IGCT geometric calibration method reduces image artifacts when uncertainties exist in system geometry. PMID:27375313

  19. CEILINEX 2015: Validation of calibration methods during the ceilometer inter-comparison

    NASA Astrophysics Data System (ADS)

    Hervo, Maxime

    2016-04-01

    In Europe, more than 700 ceilometers are measuring continuously. These instruments can be used for many applications such as detection of cloud base and aerosol layers height, aerosol profiling or for fog now-casting. However, from different manufacturers exist and the results can vary extensively from one type to another. During the CeiLinEx2015 campaign (Ceilometer Performance Experiment at Lindenberg 2015), 6 types of ceilometers (CL31, CL51, CHM15k, CHM15kx CS135 and LD40) were measuring simultaneously at Lindenberg (Germany) from June to September 2015. Each type was represented by two instruments in order to assess the instrument-to-instrument variability. A companion contribution by Mattis et al. presents an overview of the campaign. The monitoring of the temporal and spatial evolution of aerosol layers like the volcanic ash, is crucial to compare measurements from different sites. Therefore, all instruments need to be calibrated in order to provide consistent results. This contribution will focus on the validation and the comparison of state-of-the-art calibration methods. The calibration methods tested were are the cloud calibration (O'Connor et al., 2004) and the Rayleigh calibration Method (Wiegner and Geiß, 2012). Both methods can be applied without on-site intervention and are thus suitable for automatic networks. Operational automated algorithms based on these methods were developed in the framework of the TOPROF project (ESSEM COST Action ES1303)..The Cloud calibration was found more appropriate for analog instruments with analog signal detection measuring at around 905nm (Vaisala CL51 and CL31 and Campbell Scientific CS135). The Rayleigh was more suitable for photon-counting systems measuring at 1064nm (CHM15k, CHM15kx). For the first time, these methods were tested simultaneously on different instrument types and compared amongst each other. For a dust event occurred on the 13/08/2015, the attenuated backscatter coefficient difference amongst all

  20. A method for the inter-calibration of a matrix of sensors

    NASA Astrophysics Data System (ADS)

    Donetti, M.; Garelli, E.; Marchetto, F.; Boriano, A.; Bourhaleb, F.; Cirio, R.; Cornelius, I.; Giordanengo, S.; La Rosa, A.; Nastasi, U.; Peroni, C.

    2006-02-01

    We present a quick and easy method for the calibration of a matrix of sensors. The algorithm is based on a three-step irradiation procedure which relies only on the constancy of the delivered fluence at each step. With this method the gain of each sensor is derived relative to a reference detector. The algorithm has been applied to a matrix of (32 × 32) ionization chambers. After the calibration coefficients have been applied, by comparing the response of the matrix of chambers to a reference detector over a large field, we determined that the fluence measurement of individual chambers is better than 0.7%. The algorithm solves the cumbersome problem of the relative gain calibration of a matrix of a large number of sensors.

  1. A method for the inter-calibration of a matrix of sensors.

    PubMed

    Donetti, M; Garelli, E; Marchetto, F; Boriano, A; Bourhaleb, F; Cirio, R; Cornelius, I; Giordanengo, S; La Rosa, A; Nastasi, U; Peroni, C

    2006-02-07

    We present a quick and easy method for the calibration of a matrix of sensors. The algorithm is based on a three-step irradiation procedure which relies only on the constancy of the delivered fluence at each step. With this method the gain of each sensor is derived relative to a reference detector. The algorithm has been applied to a matrix of (32 x 32) ionization chambers. After the calibration coefficients have been applied, by comparing the response of the matrix of chambers to a reference detector over a large field, we determined that the fluence measurement of individual chambers is better than 0.7%. The algorithm solves the cumbersome problem of the relative gain calibration of a matrix of a large number of sensors.

  2. A Calibration Method for Nanowire Biosensors to Suppress Device-to-device Variation

    PubMed Central

    Ishikawa, Fumiaki N.; Curreli, Marco; Chang, Hsiao-Kang; Chen, Po-Chiang; Zhang, Rui; Cote, Richard J.; Thompson, Mark E.; Zhou, Chongwu

    2009-01-01

    Nanowire/nanotube biosensors have stimulated significant interest; however the inevitable device-to-device variation in the biosensor performance remains a great challenge. We have developed an analytical method to calibrate nanowire biosensor responses that can suppress the device-to-device variation in sensing response significantly. The method is based on our discovery of a strong correlation between the biosensor gate dependence (dIds/dVg) and the absolute response (absolute change in current, ΔI). In2O3 nanowire based biosensors for streptavidin detection were used as the model system. Studying the liquid gate effect and ionic concentration dependence of strepavidin sensing indicates that electrostatic interaction is the dominant mechanism for sensing response. Based on this sensing mechanism and transistor physics, a linear correlation between the absolute sensor response (ΔI) and the gate dependence (dIds/dVg) is predicted and confirmed experimentally. Using this correlation, a calibration method was developed where the absolute response is divided by dIds/dVg for each device, and the calibrated responses from different devices behaved almost identically. Compared to the common normalization method (normalization of the conductance/resistance/current by the initial value), this calibration method was proved advantageous using a conventional transistor model. The method presented here substantially suppresses device-to-device variation, allowing the use of nanosensors in large arrays. PMID:19921812

  3. A Novel Nonintrusive Method to Resolve the Thermal Dome Effect of Pyranometers: Radiometric Calibration and Implications

    NASA Technical Reports Server (NTRS)

    Ji. Q.; Tsay, S.-C.; Lau, K. M.; Hansell, R. A.; Butler, J. J.; Cooper, J. W.

    2011-01-01

    Traditionally the calibration equation for pyranometers assumes that the measured solar irradiance is solely proportional to the thermopile s output voltage; therefore, only a single calibration factor is derived. This causes additional measurement uncertainties because it does not capture sufficient information to correctly account for a pyranometer s thermal effect. In our updated calibration equation, temperatures from the pyranometer's dome and case are incorporated to describe the instrument's thermal behavior, and a new set of calibration constants are determined, thereby reducing measurement uncertainties. In this paper, we demonstrate why a pyranometer's uncertainty using the traditional calibration equation is always larger than a few percent, but with the new approach can become much less than 1% after the thermal issue is resolved. The highlighted calibration results are based on NIST traceable light sources under controlled laboratory conditions. The significance of the new approach lends itself to not only avoiding the uncertainty caused by a pyranometer's thermal effect but also the opportunity to better isolate and characterize other instrumental artifacts, such as angular response and nonlinearity of the thermopile, to further reduce additional uncertainties. We also discuss some of the implications, including an example of how the thermal issue can potentially impact climate studies by evaluating aerosol s direct radiative effect using field measurements with and without considering the pyranometer s thermal effect. The results of radiative transfer model simulation show that a pyranometer s thermal effect on solar irradiance measurements at the surface can be translated into a significant alteration of the calculated distribution of solar energy inside the column atmosphere.

  4. 10 CFR 2.705 - Discovery-additional methods.

    Code of Federal Regulations, 2012 CFR

    2012-01-01

    ... 10 Energy 1 2012-01-01 2012-01-01 false Discovery-additional methods. 2.705 Section 2.705 Energy NUCLEAR REGULATORY COMMISSION RULES OF PRACTICE FOR DOMESTIC LICENSING PROCEEDINGS AND ISSUANCE OF ORDERS... 147 and 181 of the Atomic Energy Act of 1954, as amended, the presiding officer may issue an...

  5. 10 CFR 2.705 - Discovery-additional methods.

    Code of Federal Regulations, 2013 CFR

    2013-01-01

    ... 10 Energy 1 2013-01-01 2013-01-01 false Discovery-additional methods. 2.705 Section 2.705 Energy NUCLEAR REGULATORY COMMISSION AGENCY RULES OF PRACTICE AND PROCEDURE Rules for Formal Adjudications § 2... 147 and 181 of the Atomic Energy Act of 1954, as amended, the presiding officer may issue an...

  6. 10 CFR 2.705 - Discovery-additional methods.

    Code of Federal Regulations, 2011 CFR

    2011-01-01

    ... 10 Energy 1 2011-01-01 2011-01-01 false Discovery-additional methods. 2.705 Section 2.705 Energy NUCLEAR REGULATORY COMMISSION RULES OF PRACTICE FOR DOMESTIC LICENSING PROCEEDINGS AND ISSUANCE OF ORDERS... 147 and 181 of the Atomic Energy Act of 1954, as amended, the presiding officer may issue an...

  7. 10 CFR 2.705 - Discovery-additional methods.

    Code of Federal Regulations, 2014 CFR

    2014-01-01

    ... 10 Energy 1 2014-01-01 2014-01-01 false Discovery-additional methods. 2.705 Section 2.705 Energy NUCLEAR REGULATORY COMMISSION AGENCY RULES OF PRACTICE AND PROCEDURE Rules for Formal Adjudications § 2... 147 and 181 of the Atomic Energy Act of 1954, as amended, the presiding officer may issue an...

  8. The pseudo-Thellier palaeointensity method: new calibration and uncertainty estimates

    NASA Astrophysics Data System (ADS)

    Paterson, Greig A.; Heslop, David; Pan, Yongxin

    2016-12-01

    Non-heating palaeointensity methods are a vital tool to explore magnetic field strength variations recorded by thermally sensitive materials of both terrestrial and extraterrestrial origin. One such method is the calibrated pseudo-Thellier method in which a specimen's natural remanent magnetization is alternating field demagnetized and replaced with a laboratory induced anhysteretic remanent magnetization (as an analogue of a thermoremanent magnetization, TRM). Using a set of 56 volcanic specimens given laboratory TRMs in fields of 10-130 μT, we refine the calibration of the pseudo-Thellier method and better define the uncertainty associated with its palaeointensity estimates. Our new calibration, obtained from 32 selected specimens, resolves the issue of non-zero intercept, which is theoretically predicted, but not satisfied by any previous calibration. The range of individual specimen calibration factors, however, is relatively large, but consistent with the variability expected for SD magnetite. We explore a number of rock magnetic parameters in an attempt to identify selection thresholds for reducing the calibration scatter, but fail to find a suitable choice. We infer that our careful selection process, which incorporates more statistics then previous studies, may be largely screening out any strong rock magnetic dependence. Some subtle grain size or mineralogical dependencies, however, remain after selection, but cannot be discerned from the scatter expected for grain size variability of SD magnetite. As a consequence of the variability in the calibration factor, the uncertainty associated with pseudo-Thellier results is much larger than previously indicated. The scatter of the calibration is ˜25 per cent of the mean value, which implies that, when combined with the scatter of results typically obtained from a single site, the uncertainty of averaged pseudo-Thellier results will always be >25 per cent. As such, pseudo-Thellier results should be

  9. An in-situ calibration method and the effects on stimulus frequency otoacoustic emissions

    PubMed Central

    2014-01-01

    Background The interference between the incoming sound wave and the acoustic energy reflected by the tympanic membrane (TM) forms a standing wave in human ear canals. The existence of standing waves causes various problems when measuring otoacoustic emissions (OAEs) that are soft sounds closely related with the functional status of the inner ear. The purpose of this study was to propose an in-situ calibration method to overcome the standing-wave problem and to improve the accuracy of OAE measurements. Methods In this study, the sound pressure level (SPL) at the TM was indirectly estimated by measuring the SPL at the entrance of the ear canal and the acoustic characteristics of the earphone system, so that sound energy entering the middle ear could be controlled more precisely. Then an in-situ calibration method based on the estimated TM SPL was proposed to control the stimulus level when measuring the stimulus frequency otoacoustic emissions (SFOAEs) evoked by swept tones. The results of swept-tone SFOAEs with the in-situ calibration were compared with two other calibration methods currently used in the clinic. Results Our results showed that the estimate of the SPL at the TM was rather successful with the maximal error less than 3.2 dB across all the six subjects. With the high definition OAE spectra achieved by using swept tones, it was found that the calibration methods currently used in the clinic might over-compensate the sound energy delivered to the middle ear around standing-wave frequencies and the SFOAE amplitude could be elevated by more than 7 dB as a consequence. In contrast, the in-situ calibration did not suffer from the standing-wave problem and the results could reflect the functional status of the inner ear more truthfully. Conclusions This study suggests that calibration methods currently used in the clinic may produce unreliable results. The in-situ calibration based on the estimated TM SPL could avoid the standing-wave problem and might be

  10. An Improved Calibration Method for Hydrazine Monitors for the United States Air Force

    SciTech Connect

    Korsah, K

    2003-07-07

    This report documents the results of Phase 1 of the ''Air Force Hydrazine Detector Characterization and Calibration Project''. A method for calibrating model MDA 7100 hydrazine detectors in the United States Air Force (AF) inventory has been developed. The calibration system consists of a Kintek 491 reference gas generation system, a humidifier/mixer system which combines the dry reference hydrazine gas with humidified diluent or carrier gas to generate the required humidified reference for calibrations, and a gas sampling interface. The Kintek reference gas generation system itself is periodically calibrated using an ORNL-constructed coulometric titration system to verify the hydrazine concentration of the sample atmosphere in the interface module. The Kintek reference gas is then used to calibrate the hydrazine monitors. Thus, coulometric titration is only used to periodically assess the performance of the Kintek reference gas generation system, and is not required for hydrazine monitor calibrations. One advantage of using coulometric titration for verifying the concentration of the reference gas is that it is a primary standard (if used for simple solutions), thereby guaranteeing, in principle, that measurements will be traceable to SI units (i.e., to the mole). The effect of humidity of the reference gas was characterized by using the results of concentrations determined by coulometric titration to develop a humidity correction graph for the Kintek 491 reference gas generation system. Using this calibration method, calibration uncertainty has been reduced by 50% compared to the current method used to calibrate hydrazine monitors in the Air Force inventory and calibration time has also been reduced by more than 20%. Significant findings from studies documented in this report are the following: (1) The Kintek 491 reference gas generation system (generator, humidifier and interface module) can be used to calibrate hydrazine detectors. (2) The Kintek system output

  11. An improved method for calibrating the gantry angles of linear accelerators.

    PubMed

    Higgins, Kyle; Treas, Jared; Jones, Andrew; Fallahian, Naz Afarin; Simpson, David

    2013-11-01

    Linear particle accelerators (linacs) are widely used in radiotherapy procedures; therefore, accurate calibrations of gantry angles must be performed to prevent the exposure of healthy tissue to excessive radiation. One of the common methods for calibrating these angles is the spirit level method. In this study, a new technique for calibrating the gantry angle of a linear accelerator was examined. A cubic phantom was constructed of Styrofoam with small lead balls, embedded at specific locations in this foam block. Several x-ray images were taken of this phantom at various gantry angles using an electronic portal imaging device on the linac. The deviation of the gantry angles were determined by analyzing the images using a customized computer program written in ImageJ (National Institutes of Health). Gantry angles of 0, 90, 180, and 270 degrees were chosen and the results of both calibration methods were compared for each of these angles. The results revealed that the image method was more precise than the spirit level method. For the image method, the average of the measured values for the selected angles of 0, 90, 180, and 270 degrees were found to be -0.086 ± 0.011, 90.018 ± 0.011, 180.178 ± 0.015, and 269.972 ± 0.006 degrees, respectively. The corresponding average values using the spirit level method were 0.2 ± 0.03, 90.2 ± 0.04, 180.1 ± 0.01, and 269.9 ± 0.05 degrees, respectively. Based on these findings, the new method was shown to be a reliable technique for calibrating the gantry angle.

  12. Virtual interferometer calibration method of a non-null interferometer for freeform surface measurements.

    PubMed

    Hao, Qun; Wang, Shaopu; Hu, Yao; Cheng, Hanglin; Chen, Meng; Li, Tengfei

    2016-12-10

    Non-null interferometry is a prospective method for aspheric and freeform surface measurements because of its potential versatility over null interferometry. Due to the existence of the retrace error, non-null testing usually has the drawbacks of low accuracy and inconvenient alignment of the surface under test (SUT). In this paper, a calibration method for freeform SUT in a non-null interferometer is introduced. This method calibrates the position and attitude of the SUT in a virtual interferometer with an optimization algorithm, and no accurate adjustment mechanism is required to adjust the SUT in the real interferometer. An application of this method in the digital moiré interferometric technique (DMIT) is presented. The retrace error and the influence of the alignment error can be removed after calibration. Experimental results indicate that with this calibration method, the DMIT can achieve a measurement repeatability of λ/20 (PV) when the freeform SUT has obvious alignment error. This is adequate for most aspheric and freeform surface measurements, especially when the vibration cannot be isolated completely and the freeform SUT is not easy to align.

  13. Research on position calibration method in infrared scanning temperature measurement system of rotary kiln

    NASA Astrophysics Data System (ADS)

    Dai, Shao-sheng; You, Chang-hui; Guo, Zhong-yuan; Cheng, Ya-jun; Yu, Liang-bing

    2016-11-01

    Aiming at the large error in the equal-interval locating method, a precise position calibration method is proposed. The proposed method improves the location measurement accuracy by introducing some feature temperature points to divide the rotary kiln into several segments, then the equal-interval locating method was applied to each segment, ultimately, a position calibration data more closing to the actual situation was got. The feature temperature points can be selected from the temperature points of kiln tyres or the highest temperature point and so on. Taking the practical application into consideration, the best result is obtained, when four feature temperature points was introduced to divide the rotary kiln into five segments. The experiment result shows that compared with the equal-interval method, the accuracy of the proposed method has raised about 5.6 times when four feature temperature points is used.

  14. Calibration method for the lidar-observed stratospheric depolarization ratio in the presence of liquid aerosol particles.

    PubMed

    Adachi, H; Shibata, T; Iwasaka, Y; Fujiwara, M

    2001-12-20

    A fine calibration of the depolarization ratio is required for a detailed interpretation of lidar-observed polar stratospheric clouds. We propose a procedure for analyzing data by using atmospheric depolarization lidar. The method is based on a plot of deltaT versus (1 - RT(-1)), where deltaT is the total depolarization ratio and RT is the total backscattering ratio. Assuming that there are only spherical particles in some altitude ranges of the lidar data, the characteristics of the plot of deltaT versus (1 - RT(-1)) lead to a simple but effective calibration method for deltaT. Additionally, the depolarization of air molecules deltam can be determined in the process of deltaT calibration. We compared determined values with theoretically calculated values for the depolarization of air to test the proposed method. The deltam value was calculated from the lidar data acquired at Ny-Alesund (79 degrees N, 12 degrees E), Svalbard in winter 1994-1995. When only sulfate aerosols were present on 24 December 1994, deltam was 0.46 +/- 0.35%. When the particles consisted of sulfate aerosols and spherical particles of polar stratospheric clouds on 4 January 1995, deltam was 0.45 +/- 0.07%. Both deltam values were in good agreement with the theoretically calculated value, 0.50 +/- 0.03%.

  15. Precision calibration method for binocular vision measurement systems based on arbitrary translations and 3D-connection information

    NASA Astrophysics Data System (ADS)

    Yang, Jinghao; Jia, Zhenyuan; Liu, Wei; Fan, Chaonan; Xu, Pengtao; Wang, Fuji; Liu, Yang

    2016-10-01

    Binocular vision systems play an important role in computer vision, and high-precision system calibration is a necessary and indispensable process. In this paper, an improved calibration method for binocular stereo vision measurement systems based on arbitrary translations and 3D-connection information is proposed. First, a new method for calibrating the intrinsic parameters of binocular vision system based on two translations with an arbitrary angle difference is presented, which reduces the effect of the deviation of the motion actuator on calibration accuracy. This method is simpler and more accurate than existing active-vision calibration methods and can provide a better initial value for the determination of extrinsic parameters. Second, a 3D-connection calibration and optimization method is developed that links the information of the calibration target in different positions, further improving the accuracy of the system calibration. Calibration experiments show that the calibration error can be reduced to 0.09%, outperforming traditional methods for the experiments of this study.

  16. Method and system for calibrating acquired spectra for use in spectral analysis

    DOEpatents

    Reber, Edward L.; Rohde, Kenneth W.; Blackwood, Larry G.

    2010-09-14

    A method for calibrating acquired spectra for use in spectral analysis includes performing Gaussian peak fitting to spectra acquired by a plurality of NaI detectors to define peak regions. A Na and annihilation doublet may be located among the peak regions. A predetermined energy level may be applied to one of the peaks in the doublet and a location of a hydrogen peak may be predicted based on the location of at least one of the peaks of the doublet. Control systems for calibrating spectra are also disclosed.

  17. Energy Calibration of the BaBar EMC Using the Pi0 Invariant Mass Method

    SciTech Connect

    Tanner, David J.; /Manchester U.

    2007-04-06

    The BaBar electromagnetic calorimeter energy calibration method was compared with the local and global peak iteration procedures, of Crystal Barrel and CLEO-II. An investigation was made of the possibility of {Upsilon}(4S) background reduction which could lead to increased statistics over a shorter time interval, for efficient calibration runs. The BaBar software package was used with unreconstructed data to study the energy response of the calorimeter, by utilizing the {pi}{sup 0} mass constraint on pairs of photon clusters.

  18. A new time calibration method for switched-capacitor-array-based waveform samplers

    DOE PAGES

    Kim, H.; Chen, C. -T.; Eclov, N.; ...

    2014-08-24

    Here we have developed a new time calibration method for the DRS4 waveform sampler that enables us to precisely measure the non-uniform sampling interval inherent in the switched-capacitor cells of the DRS4. The method uses the proportionality between the differential amplitude and sampling interval of adjacent switched-capacitor cells responding to a sawtooth-shape pulse. In the experiment, a sawtooth-shape pulse with a 40 ns period generated by a Tektronix AWG7102 is fed to a DRS4 evaluation board for calibrating the sampling intervals of all 1024 cells individually. The electronic time resolution of the DRS4 evaluation board with the new time calibrationmore » is measured to be ~2.4 ps RMS by using two simultaneous Gaussian pulses with 2.35 ns full-width at half-maximum and applying a Gaussian fit. The time resolution dependencies on the time difference with the new time calibration are measured and compared to results obtained by another method. Ultimately, the new method could be applicable for other switched-capacitor-array technology-based waveform samplers for precise time calibration.« less

  19. An Accurate Non-Cooperative Method for Measuring Textureless Spherical Target Based on Calibrated Lasers

    PubMed Central

    Wang, Fei; Dong, Hang; Chen, Yanan; Zheng, Nanning

    2016-01-01

    Strong demands for accurate non-cooperative target measurement have been arising recently for the tasks of assembling and capturing. Spherical objects are one of the most common targets in these applications. However, the performance of the traditional vision-based reconstruction method was limited for practical use when handling poorly-textured targets. In this paper, we propose a novel multi-sensor fusion system for measuring and reconstructing textureless non-cooperative spherical targets. Our system consists of four simple lasers and a visual camera. This paper presents a complete framework of estimating the geometric parameters of textureless spherical targets: (1) an approach to calibrate the extrinsic parameters between a camera and simple lasers; and (2) a method to reconstruct the 3D position of the laser spots on the target surface and achieve the refined results via an optimized scheme. The experiment results show that our proposed calibration method can obtain a fine calibration result, which is comparable to the state-of-the-art LRF-based methods, and our calibrated system can estimate the geometric parameters with high accuracy in real time. PMID:27941705

  20. Pyrgeometer Calibration for DOE-Atmospheric System Research Program Using NREL Method (Presentation)

    SciTech Connect

    Reda, I.; Stoffel, T.

    2010-03-15

    Presented at the DOE-Atmospheric System Research Program, Science Team Meeting, 15-19 March 2010, Bethesda, Maryland. The presentation: Pyrgeometer Calibration for DOE-Atmospheric System Research program using NREL Method - was presented by Ibrahim Reda and Tom Stoffel on March 15, 2010 at the 2010 ASR Science Team Meeting. March 15-19, 2010, Bethesda, Maryland.

  1. An Accurate Non-Cooperative Method for Measuring Textureless Spherical Target Based on Calibrated Lasers.

    PubMed

    Wang, Fei; Dong, Hang; Chen, Yanan; Zheng, Nanning

    2016-12-09

    Strong demands for accurate non-cooperative target measurement have been arising recently for the tasks of assembling and capturing. Spherical objects are one of the most common targets in these applications. However, the performance of the traditional vision-based reconstruction method was limited for practical use when handling poorly-textured targets. In this paper, we propose a novel multi-sensor fusion system for measuring and reconstructing textureless non-cooperative spherical targets. Our system consists of four simple lasers and a visual camera. This paper presents a complete framework of estimating the geometric parameters of textureless spherical targets: (1) an approach to calibrate the extrinsic parameters between a camera and simple lasers; and (2) a method to reconstruct the 3D position of the laser spots on the target surface and achieve the refined results via an optimized scheme. The experiment results show that our proposed calibration method can obtain a fine calibration result, which is comparable to the state-of-the-art LRF-based methods, and our calibrated system can estimate the geometric parameters with high accuracy in real time.

  2. A temperature calibration method for CDOM fluorescence LIF LiDAR

    NASA Astrophysics Data System (ADS)

    Chen, Peng; Mao, Zhihua; Huang, Haiqing; Bai, Yan; Wang, Tianyu

    2016-10-01

    The influence of temperature variations on the determined concentrations of dissolved organic matter (DOM) in water was investigated by laser induced fluorescence (LIF) technique in laboratory. The effect of temperature on CDOM fluorescence was investigated in freshwaters of Xixi River and in aqueous standards. The total luminescence spectra (TLS) of CDOM in several types of water samples with laser-induced fluorescence (LIF) measurements using a 405 nm wavelength excitation source were measured in the laboratory. A temperature calibration equation was derived to standardize CDOM fluorescence measurements to a specific reference temperature. Laboratory experiments with a portable LIF LiDAR showed that CDOM fluorescence intensity decreased as ambient water temperature increased. High correlation (R2=0.91) was observed between concentration of CDOM and fluorescence normalized to water Raman scattering with the temperature calibration method. The results demonstrated that temperature calibration is a necessary and important aspect of CDOM monitoring using in situ fluorescence sensors.

  3. Method and apparatus for calibrating multi-axis load cells in a dexterous robot

    NASA Technical Reports Server (NTRS)

    Wampler, II, Charles W. (Inventor); Platt, Jr., Robert J. (Inventor)

    2012-01-01

    A robotic system includes a dexterous robot having robotic joints, angle sensors adapted for measuring joint angles at a corresponding one of the joints, load cells for measuring a set of strain values imparted to a corresponding one of the load cells during a predetermined pose of the robot, and a host machine. The host machine is electrically connected to the load cells and angle sensors, and receives the joint angle values and strain values during the predetermined pose. The robot presses together mating pairs of load cells to form the poses. The host machine executes an algorithm to process the joint angles and strain values, and from the set of all calibration matrices that minimize error in force balance equations, selects the set of calibration matrices that is closest in a value to a pre-specified value. A method for calibrating the load cells via the algorithm is also provided.

  4. A curve fitting method for extrinsic camera calibration from a single image of a cylindrical object

    NASA Astrophysics Data System (ADS)

    Winkler, A. W.; Zagar, B. G.

    2013-08-01

    An important step in the process of optical steel coil quality assurance is to measure the proportions of width and radius of steel coils as well as the relative position and orientation of the camera. This work attempts to estimate these extrinsic parameters from single images by using the cylindrical coil itself as the calibration target. Therefore, an adaptive least-squares algorithm is applied to fit parametrized curves to the detected true coil outline in the acquisition. The employed model allows for strictly separating the intrinsic and the extrinsic parameters. Thus, the intrinsic camera parameters can be calibrated beforehand using available calibration software. Furthermore, a way to segment the true coil outline in the acquired images is motivated. The proposed optimization method yields highly accurate results and can be generalized even to measure other solids which cannot be characterized by the identification of simple geometric primitives.

  5. Assessment of measurement errors and dynamic calibration methods for three different tipping bucket rain gauges

    NASA Astrophysics Data System (ADS)

    Shedekar, Vinayak S.; King, Kevin W.; Fausey, Norman R.; Soboyejo, Alfred B. O.; Harmel, R. Daren; Brown, Larry C.

    2016-09-01

    Three different models of tipping bucket rain gauges (TBRs), viz. HS-TB3 (Hydrological Services Pty Ltd.), ISCO-674 (Isco, Inc.) and TR-525 (Texas Electronics, Inc.), were calibrated in the lab to quantify measurement errors across a range of rainfall intensities (5 mm·h- 1 to 250 mm·h- 1) and three different volumetric settings. Instantaneous and cumulative values of simulated rainfall were recorded at 1, 2, 5, 10 and 20-min intervals. All three TBR models showed a substantial deviation (α = 0.05) in measurements from actual rainfall depths, with increasing underestimation errors at greater rainfall intensities. Simple linear regression equations were developed for each TBR to correct the TBR readings based on measured intensities (R2 > 0.98). Additionally, two dynamic calibration techniques, viz. quadratic model (R2 > 0.7) and T vs. 1/Q model (R2 = > 0.98), were tested and found to be useful in situations when the volumetric settings of TBRs are unknown. The correction models were successfully applied to correct field-collected rainfall data from respective TBR models. The calibration parameters of correction models were found to be highly sensitive to changes in volumetric calibration of TBRs. Overall, the HS-TB3 model (with a better protected tipping bucket mechanism, and consistent measurement errors across a range of rainfall intensities) was found to be the most reliable and consistent for rainfall measurements, followed by the ISCO-674 (with susceptibility to clogging and relatively smaller measurement errors across a range of rainfall intensities) and the TR-525 (with high susceptibility to clogging and frequent changes in volumetric calibration, and highly intensity-dependent measurement errors). The study demonstrated that corrections based on dynamic and volumetric calibration can only help minimize-but not completely eliminate the measurement errors. The findings from this study will be useful for correcting field data from TBRs; and may have major

  6. Optics-Only Calibration of a Neural-Net Based Optical NDE Method for Structural Health Monitoring

    NASA Technical Reports Server (NTRS)

    Decker, Arthur J.

    2004-01-01

    A calibration process is presented that uses optical measurements alone to calibrate a neural-net based NDE method. The method itself detects small changes in the vibration mode shapes of structures. The optics-only calibration process confirms previous work that the sensitivity to vibration-amplitude changes can be as small as 10 nanometers. A more practical value in an NDE service laboratory is shown to be 50 nanometers. Both model-generated and experimental calibrations are demonstrated using two implementations of the calibration technique. The implementations are based on previously published demonstrations of the NDE method and an alternative calibration procedure that depends on comparing neural-net and point sensor measurements. The optics-only calibration method, unlike the alternative method, does not require modifications of the structure being tested or the creation of calibration objects. The calibration process can be used to test improvements in the NDE process and to develop a vibration-mode-independence of damagedetection sensitivity. The calibration effort was intended to support NASA s objective to promote safety in the operations of ground test facilities or aviation safety, in general, by allowing the detection of the gradual onset of structural changes and damage.

  7. Absolute calibration method for nanosecond-resolved, time-streaked, fiber optic light collection, spectroscopy systems

    NASA Astrophysics Data System (ADS)

    Johnston, Mark D.; Oliver, Bryan V.; Droemer, Darryl W.; Frogget, Brent; Crain, Marlon D.; Maron, Yitzhak

    2012-08-01

    This paper describes a convenient and accurate method to calibrate fast (<1 ns resolution) streaked, fiber optic light collection, spectroscopy systems. Such systems are inherently difficult to calibrate due to the lack of sufficiently intense, calibrated light sources. Such a system is used to collect spectral data on plasmas generated in electron beam diodes fielded on the RITS-6 accelerator (8-12MV, 140-200kA) at Sandia National Laboratories. On RITS, plasma light is collected through a small diameter (200 μm) optical fiber and recorded on a fast streak camera at the output of a 1 meter Czerny-Turner monochromator. For this paper, a 300 W xenon short arc lamp (Oriel Model 6258) was used as the calibration source. Since the radiance of the xenon arc varies from cathode to anode, just the area around the tip of the cathode ("hotspot") was imaged onto the fiber, to produce the highest intensity output. To compensate for chromatic aberrations, the signal was optimized at each wavelength measured. Output power was measured using 10 nm bandpass interference filters and a calibrated photodetector. These measurements give power at discrete wavelengths across the spectrum, and when linearly interpolated, provide a calibration curve for the lamp. The shape of the spectrum is determined by the collective response of the optics, monochromator, and streak tube across the spectral region of interest. The ratio of the spectral curve to the measured bandpass filter curve at each wavelength produces a correction factor (Q) curve. This curve is then applied to the experimental data and the resultant spectra are given in absolute intensity units (photons/sec/cm2/steradian/nm). Error analysis shows this method to be accurate to within +/- 20%, which represents a high level of accuracy for this type of measurement.

  8. Absolute calibration method for nanosecond-resolved, time-streaked, fiber optic light collection, spectroscopy systems.

    PubMed

    Johnston, Mark D; Oliver, Bryan V; Droemer, Darryl W; Frogget, Brent; Crain, Marlon D; Maron, Yitzhak

    2012-08-01

    This paper describes a convenient and accurate method to calibrate fast (<1 ns resolution) streaked, fiber optic light collection, spectroscopy systems. Such systems are inherently difficult to calibrate due to the lack of sufficiently intense, calibrated light sources. Such a system is used to collect spectral data on plasmas generated in electron beam diodes fielded on the RITS-6 accelerator (8-12MV, 140-200kA) at Sandia National Laboratories. On RITS, plasma light is collected through a small diameter (200 μm) optical fiber and recorded on a fast streak camera at the output of a 1 meter Czerny-Turner monochromator. For this paper, a 300 W xenon short arc lamp (Oriel Model 6258) was used as the calibration source. Since the radiance of the xenon arc varies from cathode to anode, just the area around the tip of the cathode ("hotspot") was imaged onto the fiber, to produce the highest intensity output. To compensate for chromatic aberrations, the signal was optimized at each wavelength measured. Output power was measured using 10 nm bandpass interference filters and a calibrated photodetector. These measurements give power at discrete wavelengths across the spectrum, and when linearly interpolated, provide a calibration curve for the lamp. The shape of the spectrum is determined by the collective response of the optics, monochromator, and streak tube across the spectral region of interest. The ratio of the spectral curve to the measured bandpass filter curve at each wavelength produces a correction factor (Q) curve. This curve is then applied to the experimental data and the resultant spectra are given in absolute intensity units (photons/sec/cm(2)/steradian/nm). Error analysis shows this method to be accurate to within +∕- 20%, which represents a high level of accuracy for this type of measurement.

  9. Semi-automated calibration method for modelling of mountain permafrost evolution in Switzerland

    NASA Astrophysics Data System (ADS)

    Marmy, Antoine; Rajczak, Jan; Delaloye, Reynald; Hilbich, Christin; Hoelzle, Martin; Kotlarski, Sven; Lambiel, Christophe; Noetzli, Jeannette; Phillips, Marcia; Salzmann, Nadine; Staub, Benno; Hauck, Christian

    2016-11-01

    Permafrost is a widespread phenomenon in mountainous regions of the world such as the European Alps. Many important topics such as the future evolution of permafrost related to climate change and the detection of permafrost related to potential natural hazards sites are of major concern to our society. Numerical permafrost models are the only tools which allow for the projection of the future evolution of permafrost. Due to the complexity of the processes involved and the heterogeneity of Alpine terrain, models must be carefully calibrated, and results should be compared with observations at the site (borehole) scale. However, for large-scale applications, a site-specific model calibration for a multitude of grid points would be very time-consuming. To tackle this issue, this study presents a semi-automated calibration method using the Generalized Likelihood Uncertainty Estimation (GLUE) as implemented in a 1-D soil model (CoupModel) and applies it to six permafrost sites in the Swiss Alps. We show that this semi-automated calibration method is able to accurately reproduce the main thermal condition characteristics with some limitations at sites with unique conditions such as 3-D air or water circulation, which have to be calibrated manually. The calibration obtained was used for global and regional climate model (GCM/RCM)-based long-term climate projections under the A1B climate scenario (EU-ENSEMBLES project) specifically downscaled at each borehole site. The projection shows general permafrost degradation with thawing at 10 m, even partially reaching 20 m depth by the end of the century, but with different timing among the sites and with partly considerable uncertainties due to the spread of the applied climatic forcing.

  10. New Method for the Calibration of Multi-Camera Mobile Mapping Systems

    NASA Astrophysics Data System (ADS)

    Kersting, A. P.; Habib, A.; Rau, J.

    2012-07-01

    Mobile Mapping Systems (MMS) allow for fast and cost-effective collection of geo-spatial information. Such systems integrate a set of imaging sensors and a position and orientation system (POS), which entails GPS and INS units. System calibration is a crucial process to ensure the attainment of the expected accuracy of such systems. It involves the calibration of the individual sensors as well as the calibration of the mounting parameters relating the system components. The mounting parameters of multi-camera MMS include two sets of relative orientation parameters (ROP): the lever arm offsets and the boresight angles relating the cameras and the IMU body frame and the ROP among the cameras (in the absence of GPS/INS data). In this paper, a novel single-step calibration method, which has the ability of estimating these two sets of ROP, is devised. Besides the ability to estimate the ROP among the cameras, the proposed method can use such parameters as prior information in the ISO procedure. The implemented procedure consists of an integrated sensor orientation (ISO) where the GPS/INS-derived position and orientation and the system mounting parameters are directly incorporated in the collinearity equations. The concept of modified collinearity equations has been used by few authors for single-camera systems. In this paper, a new modification to the collinearity equations for GPS/INS-assisted multicamera systems is introduced. Experimental results using a real dataset demonstrate the feasibility of the proposed method.

  11. A new calibration method between an optical sensor and a rotating platform in turbine blade inspection

    NASA Astrophysics Data System (ADS)

    Li, Wen-Long; Wu, An; Li, Zi-Cheng; Zhang, Gang; Yu, Wen-Yong

    2017-03-01

    Thin-walled turbine blades with complex features are a critical part of an aviation engine, and a small change in their geometric shape can erode the performance of the aviation engine. Inspecting the blade with an optical device is a promising technique. One key task involved is the calibration of the optical sensor with the rotating platform. This paper presents a novel calibration method for the optical inspection of the blade. Three target spheres are measured by a high-precision laser tracker and an optical sensor. The positions of the sphere centers are used to build a coordinate system and an approximated plane. Following that, the rotation axis and the rotation center of the rotating platform are easily calculated. According to a direction vector from the measured stripe, the transformation parameters between the optical sensor and rotating platform are further calculated. This calibration method is simple to carry out, and it guarantees that all the measured points are represented in the same coordinate system for subsequent parameter extraction and profile error evaluation of the blade surface. The experiments demonstrate the feasibility of the proposed method, and it found that the measurement error after calibration is within 0.02 mm.

  12. Error Analysis and Calibration Method of a Multiple Field-of-View Navigation System

    PubMed Central

    Shi, Shuai; Zhao, Kaichun; You, Zheng; Ouyang, Chenguang; Cao, Yongkui; Wang, Zhenzhou

    2017-01-01

    The Multiple Field-of-view Navigation System (MFNS) is a spacecraft subsystem built to realize the autonomous navigation of the Spacecraft Inside Tiangong Space Station. This paper introduces the basics of the MFNS, including its architecture, mathematical model and analysis, and numerical simulation of system errors. According to the performance requirement of the MFNS, the calibration of both intrinsic and extrinsic parameters of the system is assumed to be essential and pivotal. Hence, a novel method based on the geometrical constraints in object space, called checkerboard-fixed post-processing calibration (CPC), is proposed to solve the problem of simultaneously obtaining the intrinsic parameters of the cameras integrated in the MFNS and the transformation between the MFNS coordinate and the cameras’ coordinates. This method utilizes a two-axis turntable and a prior alignment of the coordinates is needed. Theoretical derivation and practical operation of the CPC method are introduced. The calibration experiment results of the MFNS indicate that the extrinsic parameter accuracy of the CPC reaches 0.1° for each Euler angle and 0.6 mm for each position vector component (1σ). A navigation experiment verifies the calibration result and the performance of the MFNS. The MFNS is found to work properly, and the accuracy of the position vector components and Euler angle reaches 1.82 mm and 0.17° (1σ) respectively. The basic mechanism of the MFNS may be utilized as a reference for the design and analysis of multiple-camera systems. Moreover, the calibration method proposed has practical value for its convenience for use and potential for integration into a toolkit. PMID:28327538

  13. Characterizing a switching reagent ion chemical ionization high resolution time of flight mass spectrometer: Standard additions, External calibrations, and Inlet response during SOAS

    NASA Astrophysics Data System (ADS)

    Brophy, P.; Farmer, D.

    2013-12-01

    A high-resolution time of flight chemical ionization mass spectrometer (HRToF-CIMS) with switching reagent ion source and low pressure, gas-phase inlet was deployed during the 2013 Southern Oxidant and Aerosol Study (SOAS) in Brent, Alabama. Acetate chemistry was employed for the detection of small acids and iodine chemistry for the detection of peroxy acids. Switching between the two ion sources was found to be possible on less than ten minute time scales with minimal artifacts observed. Online calibrations for formic acid on both the acetate and iodine sources were performed every hour using both standard addition techniques as well as external standard calibrations; offline formic acid calibrations were also conducted. Inlet responses were investigated though a number of experiments finding that the inlet has minimal hysteresis and rapid response times.

  14. Empirical mode decomposition-adaptive least squares method for dynamic calibration of pressure sensors

    NASA Astrophysics Data System (ADS)

    Yao, Zhenjian; Wang, Zhongyu; Yi-Lin Forrest, Jeffrey; Wang, Qiyue; Lv, Jing

    2017-04-01

    In this paper, an approach combining empirical mode decomposition (EMD) with adaptive least squares (ALS) is proposed to improve the dynamic calibration accuracy of pressure sensors. With EMD, the original output of the sensor can be represented as sums of zero-mean amplitude modulation frequency modulation components. By identifying and excluding those components involved in noises, the noise-free output could be reconstructed with the useful frequency modulation ones. Then the least squares method is iteratively performed to estimate the optimal order and parameters of the mathematical model. The dynamic characteristic parameters of the sensor can be derived from the model in both time and frequency domains. A series of shock tube calibration tests are carried out to validate the performance of this method. Experimental results show that the proposed method works well in reducing the influence of noise and yields an appropriate mathematical model. Furthermore, comparative experiments also demonstrate the superiority of the proposed method over the existing ones.

  15. Interferometric method for phase calibration in liquid crystal spatial light modulators using a self-generated diffraction-grating.

    PubMed

    Fuentes, José Luis Martínez; Fernández, Enrique J; Prieto, Pedro M; Artal, Pablo

    2016-06-27

    An auto-referenced interferometric method for calibrating phase modulation of parallel-aligned liquid crystal (PAL) spatial light modulators (SLM) is described. The method is experimentally straightforward, robust, and requires solely of a collimated beam, with no need of additional optics. This method uses the SLM itself to create a tilted plane wave and a reference wave which mutually interfere. These waves are codified by means of a binary diffraction grating and a uniformly distributed gray level area (piston) into the SLM surface. Phase shift for each gray level addressed to the piston section can then be evaluated. Phase modulation on the SLM can also be retrieved with the proposed method over spatially resolved portions of the surface. Phase information obtained with this novel method is compared to other well established calibration procedures, requiring extra elements and more elaborated optical set-ups. The results show a good agreement with previous methods. The advantages of the new method include high mechanical stability, faster performance, and a significantly easier practical implementation.

  16. Development and evaluation of a method of calibrating medical displays based on fixed adaptation

    SciTech Connect

    Sund, Patrik Månsson, Lars Gunnar; Båth, Magnus

    2015-04-15

    Purpose: The purpose of this work was to develop and evaluate a new method for calibration of medical displays that includes the effect of fixed adaptation and by using equipment and luminance levels typical for a modern radiology department. Methods: Low contrast sinusoidal test patterns were derived at nine luminance levels from 2 to 600 cd/m{sup 2} and used in a two alternative forced choice observer study, where the adaptation level was fixed at the logarithmic average of 35 cd/m{sup 2}. The contrast sensitivity at each luminance level was derived by establishing a linear relationship between the ten pattern contrast levels used at every luminance level and a detectability index (d′) calculated from the fraction of correct responses. A Gaussian function was fitted to the data and normalized to the adaptation level. The corresponding equation was used in a display calibration method that included the grayscale standard display function (GSDF) but compensated for fixed adaptation. In the evaluation study, the contrast of circular objects with a fixed pixel contrast was displayed using both calibration methods and was rated on a five-grade scale. Results were calculated using a visual grading characteristics method. Error estimations in both observer studies were derived using a bootstrap method. Results: The contrast sensitivities for the darkest and brightest patterns compared to the contrast sensitivity at the adaptation luminance were 37% and 56%, respectively. The obtained Gaussian fit corresponded well with similar studies. The evaluation study showed a higher degree of equally distributed contrast throughout the luminance range with the calibration method compensated for fixed adaptation than for the GSDF. The two lowest scores for the GSDF were obtained for the darkest and brightest patterns. These scores were significantly lower than the lowest score obtained for the compensated GSDF. For the GSDF, the scores for all luminance levels were statistically

  17. Hybrid PSO-ASVR-based method for data fitting in the calibration of infrared radiometer.

    PubMed

    Yang, Sen; Li, Chengwei

    2016-06-01

    The present paper describes a hybrid particle swarm optimization-adaptive support vector regression (PSO-ASVR)-based method for data fitting in the calibration of infrared radiometer. The proposed hybrid PSO-ASVR-based method is based on PSO in combination with Adaptive Processing and Support Vector Regression (SVR). The optimization technique involves setting parameters in the ASVR fitting procedure, which significantly improves the fitting accuracy. However, its use in the calibration of infrared radiometer has not yet been widely explored. Bearing this in mind, the PSO-ASVR-based method, which is based on the statistical learning theory, is successfully used here to get the relationship between the radiation of a standard source and the response of an infrared radiometer. Main advantages of this method are the flexible adjustment mechanism in data processing and the optimization mechanism in a kernel parameter setting of SVR. Numerical examples and applications to the calibration of infrared radiometer are performed to verify the performance of PSO-ASVR-based method compared to conventional data fitting methods.

  18. Hybrid PSO-ASVR-based method for data fitting in the calibration of infrared radiometer

    NASA Astrophysics Data System (ADS)

    Yang, Sen; Li, Chengwei

    2016-06-01

    The present paper describes a hybrid particle swarm optimization-adaptive support vector regression (PSO-ASVR)-based method for data fitting in the calibration of infrared radiometer. The proposed hybrid PSO-ASVR-based method is based on PSO in combination with Adaptive Processing and Support Vector Regression (SVR). The optimization technique involves setting parameters in the ASVR fitting procedure, which significantly improves the fitting accuracy. However, its use in the calibration of infrared radiometer has not yet been widely explored. Bearing this in mind, the PSO-ASVR-based method, which is based on the statistical learning theory, is successfully used here to get the relationship between the radiation of a standard source and the response of an infrared radiometer. Main advantages of this method are the flexible adjustment mechanism in data processing and the optimization mechanism in a kernel parameter setting of SVR. Numerical examples and applications to the calibration of infrared radiometer are performed to verify the performance of PSO-ASVR-based method compared to conventional data fitting methods.

  19. Transient thermal dissipation method of xylem sap flow measurement: multi-species calibration and field evaluation.

    PubMed

    Isarangkool Na Ayutthaya, S; Do, F C; Pannengpetch, K; Junjittakarn, J; Maeght, J-L; Rocheteau, A; Cochard, H

    2010-01-01

    The transient thermal dissipation (TTD) method developed by Do and Rocheteau (2002b) is a close evolution of the original constant thermal dissipation (CTD) method of Granier (1985). The TTD method has the advantage of limiting the influence of passive natural temperature gradients and of yielding more stable zero-flux references at night. By analogy with the CTD method, the transient method was first calibrated on synthetic porous material (sawdust) on the assumption that the relationship was independent of the woody species. Here, our concern was to test the latter hypothesis with a 10-min heating time in three tropical species: Hevea brasiliensis Müll. Arg., Mangifera indica L. and Citrus maxima Merr. A complementary objective was to compare the field estimates of daily transpiration for mature rubber trees with estimates based on a simplified soil water balance in the dry season. The calibration experiments were carried out in the laboratory on cut stems using an HPFM device and gravimetric control of water flow up to 5 L dm(-2) h(-1). Nineteen response curves were assessed on fully conductive xylem, combining 11 cut stems and two probes. The field evaluation comprised five periods from November 2007 to February 2008. Estimates of daily transpiration from the measurement of sap flow were based on the 41 sensors set up on 11 trees. Soil water depletion was monitored by neutron probe and 12 access tubes to a depth of 1.8 m. The calibrations confirmed that the response of the transient thermal index to flow density was independent of the woody species that were tested. The best fit was a simple linear response (R(2) = 0.88, n = 276 and P < 0.0001). The previous calibration performed by Do and Rocheteau (2002b) on sawdust fell within the variability of the multi-species calibration; however, there were substantial differences with the average curve at extreme flow rates. Field comparison with soil water depletion in the dry season validated to a reasonable extent

  20. Self-referenced method for optical path difference calibration in low-coherence interferometry.

    PubMed

    Laubscher, M; Froehly, L; Karamata, B; Salathé, R P; Lasser, T

    2003-12-15

    A simple method for the calibration of optical path difference modulation in low-coherence interferometry is presented. Spectrally filtering a part of the detected interference signal results in a high-coherence signal that encodes the scan imperfections and permits their correction. The method is self-referenced in the sense that no secondary high-coherence light source is necessary. Using a spectrometer setup for spectral filtering allows for flexibility in both the choice of calibration wavelength and the maximum scan range. To demonstrate the method's usefulness, it is combined with a recently published digital spectral shaping technique to measure the thickness of a pellicle beam splitter with a white-light source.

  1. A Focusing Method in the Calibration Process of Image Sensors Based on IOFBs

    PubMed Central

    Fernández, Pedro R.; Lázaro, José L.; Gardel, Alfredo; Cano, Ángel E.; Bravo, Ignacio

    2010-01-01

    A focusing procedure in the calibration process of image sensors based on Incoherent Optical Fiber Bundles (IOFBs) is described using the information extracted from fibers. These procedures differ from any other currently known focusing method due to the non spatial in-out correspondence between fibers, which produces a natural codification of the image to transmit. Focus measuring is essential prior to carrying out calibration in order to guarantee accurate processing and decoding. Four algorithms have been developed to estimate the focus measure; two methods based on mean grey level, and the other two based on variance. In this paper, a few simple focus measures are defined and compared. Some experimental results referred to the focus measure and the accuracy of the developed methods are discussed in order to demonstrate its effectiveness. PMID:22315526

  2. Determination of intrinsic viscosity of poly(ethylene terephthalate) using infrared spectroscopy and multivariate calibration method.

    PubMed

    Silva Spinacé, M A; Lucato, M U; Ferrão, M F; Davanzo, C U; De Paoli, M-A

    2006-05-15

    A methodology was developed to determine the intrinsic viscosity of poly(ethylene terephthalate) (PET) using diffuse reflectance infrared Fourier transform spectroscopy (DRIFTS) and multivariate calibration (MVC) methods. Multivariate partial least squares calibration was applied to the spectra using mean centering and cross validation. The results were correlated to the intrinsic viscosities determined by the standard chemical method (ASTM D 4603-01) and a very good correlation for values in the range from 0.346 to 0.780dLg(-1) (relative viscosity values ca. 1.185-1.449) was observed. The spectrophotometer detector sensitivity and the humidity of the samples did not influence the results. The methodology developed is interesting because it does not produce hazardous wastes, avoids the use of time-consuming chemical methods and can rapidly predict the intrinsic viscosity of PET samples over a large range of values, which includes those of recycled materials.

  3. A novel calibration method for PD measurements in power cables and joints using capacitive couplers

    NASA Astrophysics Data System (ADS)

    Zhong, L.; Chen, G.; Xu, Y.

    2004-09-01

    Partial discharge (PD) measurements are universally accepted as a technique giving some indication of the state of the insulation in high-voltage apparatus. Cable end users are keen to adopt online PD monitoring during commissioning of systems. However, because of noisy environments and the problems of interference the conventional methods are difficult to implement. As a consequence, ultra-high frequency (UHF)/very high frequency (VHF) techniques of on-site detection of partial discharges have been developed. A technique based on capacitive coupling has received much attention. It has been demonstrated that it is possible to use the technique to monitor the partial discharge in cables, particularly in joints/terminations. However, in order to obtain quantitative information about PD, calibration is required for this technique. Existing calibration methods are difficult to implement on-site. In this paper, a novel method is proposed and compared with the conventional method on a short piece of cable. It has been shown that an individual capacitive coupler can also be accurately calibrated on-site and online using the new method, therefore it provides quantitative information about the amount of apparent discharge. In practice, this is important for electricity utilities as the quantitative information about PD can be used to determine the quality of the cable system and to decide whether the system needs to be repaired or replaced.

  4. An improved method of energy calibration for position-sensitive silicon detectors

    NASA Astrophysics Data System (ADS)

    Sun, Ming-Dao; Huang, Tian-Heng; Liu, Zhong; Ding, Bing; Yang, Hua-Bin; Zhang, Zhi-Yuan; Wang, Jian-Guo; Ma, Long; Yu, Lin; Wang, Yong-Sheng; Gan, Zai-Guo; Xiao-Hong, Zhou

    2016-04-01

    Energy calibration of resistive charge division-based position-sensitive silicon detectors is achieved by parabolic fitting in the traditional method, where the systematic variations of vertex and curvature of the parabola with energy must be considered. In this paper we extend the traditional method in order to correct the fitting function, simplify the procedure of calibration and improve the experimental data quality. Instead of a parabolic function as used in the traditional method, a new function describing the relation of position and energy is introduced. The energy resolution of the 8.088 MeV α decay of 213Rn is determined to be about 87 keV (FWHM), which is better than the result of the traditional method, 104 keV (FWHM). The improved method can be applied to the energy calibration of resistive charge division-based position-sensitive silicon detectors with various performances. Supported by ‘100 Person Project’ of the Chinese Academy of Sciences and the National Natural Science Foundation of China (11405224 and 11435014)

  5. Optical Calibration Process Developed for Neural-Network-Based Optical Nondestructive Evaluation Method

    NASA Technical Reports Server (NTRS)

    Decker, Arthur J.

    2004-01-01

    A completely optical calibration process has been developed at Glenn for calibrating a neural-network-based nondestructive evaluation (NDE) method. The NDE method itself detects very small changes in the characteristic patterns or vibration mode shapes of vibrating structures as discussed in many references. The mode shapes or characteristic patterns are recorded using television or electronic holography and change when a structure experiences, for example, cracking, debonds, or variations in fastener properties. An artificial neural network can be trained to be very sensitive to changes in the mode shapes, but quantifying or calibrating that sensitivity in a consistent, meaningful, and deliverable manner has been challenging. The standard calibration approach has been difficult to implement, where the response to damage of the trained neural network is compared with the responses of vibration-measurement sensors. In particular, the vibration-measurement sensors are intrusive, insufficiently sensitive, and not numerous enough. In response to these difficulties, a completely optical alternative to the standard calibration approach was proposed and tested successfully. Specifically, the vibration mode to be monitored for structural damage was intentionally contaminated with known amounts of another mode, and the response of the trained neural network was measured as a function of the peak-to-peak amplitude of the contaminating mode. The neural network calibration technique essentially uses the vibration mode shapes of the undamaged structure as standards against which the changed mode shapes are compared. The published response of the network can be made nearly independent of the contaminating mode, if enough vibration modes are used to train the net. The sensitivity of the neural network can be adjusted for the environment in which the test is to be conducted. The response of a neural network trained with measured vibration patterns for use on a vibration isolation

  6. A flexible tactile sensor calibration method based on an air-bearing six-dimensional force measurement platform.

    PubMed

    Huang, Bin

    2015-07-01

    A number of common issues related to the process of flexible tactile sensor calibration are discussed in this paper, and an estimate of the accuracy of classical calibration methods, as represented by a weight-pulley device, is presented. A flexible tactile sensor calibration method that is based on a six-dimensional force measurement is proposed on the basis of a theoretical analysis. A high-accuracy flexible tactile sensor calibration bench based on the air-bearing six-dimensional force measurement principle was developed to achieve a technically challenging measurement accuracy of 2% full scale (FS) for three-dimensional (3D) flexible tactile sensor calibration. The experimental results demonstrate that the accuracy of the air-bearing six-dimensional force measurement platform can reach 0.2% FS. Thus, the system satisfies the 3D flexible tactile sensor calibration requirement of 2% FS.

  7. High-precision method of binocular camera calibration with a distortion model.

    PubMed

    Li, Weimin; Shan, Siyu; Liu, Hui

    2017-03-10

    A high-precision camera calibration method for binocular stereo vision system based on a multi-view template and alternative bundle adjustment is presented in this paper. The proposed method could be achieved by taking several photos on a specially designed calibration template that has diverse encoded points in different orientations. In this paper, the method utilized the existing algorithm used for monocular camera calibration to obtain the initialization, which involves a camera model, including radial lens distortion and tangential distortion. We created a reference coordinate system based on the left camera coordinate to optimize the intrinsic parameters of left camera through alternative bundle adjustment to obtain optimal values. Then, optimal intrinsic parameters of the right camera can be obtained through alternative bundle adjustment when we create a reference coordinate system based on the right camera coordinate. We also used all intrinsic parameters that were acquired to optimize extrinsic parameters. Thus, the optimal lens distortion parameters and intrinsic and extrinsic parameters were obtained. Synthetic and real data were used to test the method. The simulation results demonstrate that the maximum mean absolute relative calibration errors are about 3.5e-6 and 1.2e-6 for the focal length and the principal point, respectively, under zero-mean Gaussian noise with 0.05 pixels standard deviation. The real result shows that the reprojection error of our model is about 0.045 pixels with the relative standard deviation of 1.0e-6 over the intrinsic parameters. The proposed method is convenient, cost-efficient, highly precise, and simple to carry out.

  8. Methods for the additive manufacturing of semiconductor and crystal materials

    SciTech Connect

    Stowe, Ashley C.; Speight, Douglas

    2016-11-22

    A method for the additive manufacturing of inorganic crystalline materials, including: physically combining a plurality of starting materials that are used to form an inorganic crystalline compound to be used as one or more of a semiconductor, scintillator, laser crystal, and optical filter; heating or melting successive regions of the combined starting materials using a directed heat source having a predetermined energy characteristic, thereby facilitating the reaction of the combined starting materials; and allowing each region of the combined starting materials to cool in a controlled manner, such that the desired inorganic crystalline compound results. The method also includes, prior to heating or melting the successive regions of the combined starting materials using the directed heat source, heating the combined starting materials to facilitate initial reaction of the combined starting materials. The method further includes translating the combined starting materials and/or the directed heat source between successive locations. The method still further includes controlling the mechanical, electrical, photonic, and/or optical properties of the inorganic crystalline compound.

  9. A novel target LOS calibration method for IR scanning sensor based on control points

    NASA Astrophysics Data System (ADS)

    Xue, Yong-Hong; An, Wei; Zhang, Yin-Sheng; Zhang, Tao

    2012-12-01

    Space based IR system uses the information of target LOS (line of sight) for target location. Recent researches show that the measuring precision of target LOS is usually determined by measuring precision of platform's position and attitude, and deformation of sensor etc. Most methods for improving target location precision are either through improving platform's position and attitude measuring precision or through calib rating the whole image obtained by IR sensor. With the development of measuring technology, it is harder to make a further improvement on the measuring precision of position and attitude of the platform and the expansion of the sensor view make calibrat ion of the whole image with a larger computation cost. In this paper, a method using control points to calibrate target LOS was proposed. Based on the analysis of the imaging process of the scanning sensor of space based IR system, this paper established a modify model of target LOS based on control points, used a bias filter to estimate the bias value of sensor boresight, and finally achieved the mission of target LOS calibrat ion. Different from the traditional calibration method of remote sensing image, the proposed method only made a correct ion on the LOS of suspicious target, but not established the accurate relationship between the all pixels and their real location, and has a similar calibration performance, but more lower computational complexity.

  10. Generic precise augmented reality guiding system and its calibration method based on 3D virtual model.

    PubMed

    Liu, Miao; Yang, Shourui; Wang, Zhangying; Huang, Shujun; Liu, Yue; Niu, Zhenqi; Zhang, Xiaoxuan; Zhu, Jigui; Zhang, Zonghua

    2016-05-30

    Augmented reality system can be applied to provide precise guidance for various kinds of manual works. The adaptability and guiding accuracy of such systems are decided by the computational model and the corresponding calibration method. In this paper, a novel type of augmented reality guiding system and the corresponding designing scheme are proposed. Guided by external positioning equipment, the proposed system can achieve high relative indication accuracy in a large working space. Meanwhile, the proposed system is realized with a digital projector and the general back projection model is derived with geometry relationship between digitized 3D model and the projector in free space. The corresponding calibration method is also designed for the proposed system to obtain the parameters of projector. To validate the proposed back projection model, the coordinate data collected by a 3D positioning equipment is used to calculate and optimize the extrinsic parameters. The final projecting indication accuracy of the system is verified with subpixel pattern projecting technique.

  11. Ozone Correction for AM0 Calibrated Solar Cells for the Aircraft Method

    NASA Technical Reports Server (NTRS)

    Snyder, David B.; Scheiman, David A.; Jenkins, Phillip P.; Lyons, Valerie J. (Technical Monitor)

    2002-01-01

    The aircraft solar cell calibration method has provided cells calibrated to space conditions for 37 years. However, it is susceptible to systematic errors due to ozone concentration in the stratosphere. The present correction procedure applies a 1% increase to the measured Isc values. High band-gap cells are more sensitive to ozone adsorbed wavelengths so it has become important to reassess the correction technique. This paper evaluates the ozone correction to be 1+{O3}sup Fo, where Fo is 29.5x10(exp-6)/d.u. for a Silicon solar cell and 42.2xl0(exp -6)/d.u. for a GaAs cell. Results will be presented for high band-gap cells. A comparison with flight data indicates that this method of correcting for the ozone density improves the uncertainty of AM0 Isc to 0.5%.

  12. Effects of sensor calibration, balancing and parametrization on the signal space separation method.

    PubMed

    Nurminen, J; Taulu, S; Okada, Y

    2008-04-07

    Signal space separation (SSS) is a novel method for processing multichannel biomagnetic data. It is useful for a variety of applications including interference suppression, movement compensation and conversion of measurements between sensor arrays. The performance of SSS has been examined mainly on a 306-channel whole-head magnetoencephalography system. To facilitate the adaptation of the method to other biomagnetometer systems, the effect of various properties of the sensor array on its performance needs to be studied. To this end, we examined the effects of gradiometer imbalance, sensor calibration errors and erroneous sensor geometry information on SSS using simulations. The results indicate that depending on the application, gradiometer balance on the level of 0.1% to 0.5% may be needed for satisfactory SSS performance. For wire-wound gradiometers, this requires very careful attention in manufacturing. Errors in calibration coefficients and geometry information were found to have less significance.

  13. A High Speed Calibration Method for Laser Positioner by Constant Velocity Scanning

    NASA Astrophysics Data System (ADS)

    Ono, Hiroyuki

    This paper describes a high speed calibration method for laser positioner by scanning work area under constant velocity. Laser positioner consists of sinusoidal laser encoder, DC motor and the controller. The encoder has diffraction grating scale and laser optics. It generates sine and cosine outputs according to the travel distance. Qualities of the scale and the optics contribute to the fluctuation of output signal amplitude, DC offset and relative phase. First, the controller collects the output data under constant velocity of 4 samples per 1 sine wave length. The parameter can be estimated by simplified discrete Fourier transform method. Calibration data are collected every quarter sine wave length over all positioner work area. The result is stored in the table then referred by the controller in real time operation. Experiment results are also reported for HDD servo track writer application.

  14. Self-Calibrating Ultrasonic Methods for In-Situ Monitoring of Fatigue Crack Progression

    SciTech Connect

    Michaels, J.E.; Mi, B.; Cobb, A.C.; Michaels, T.E.; Stobbe, D.M.

    2005-04-09

    Ultrasonic sensors permanently affixed to aluminum coupons are used to monitor progression of damage during fatigue testing with the long term goal of structural health monitoring for diagnostics and prognostics. Necessary for success are proper design of the ultrasonic testing methods, robust transducer mounting techniques, and real-time signal processing for determining the state of the structure. It is also highly desirable for the overall system to be self-calibrating with built-in diagnostics in order to detect and compensate for sensor degradation or failure. Self-calibrating ultrasonic techniques are applied for monitoring of cracks initiating and propagating from the inaccessible inner diameters of rivet holes where the transducers are mounted on the accessible specimen surface. Angle beam ultrasonic methods are utilized that are suitable for detecting small defects in critical local regions of high stress. Results are presented for aluminum coupons subjected to low cycle fatigue and demonstrate ultrasonic tracking of crack growth.

  15. Self-Calibrating Ultrasonic Methods for In-Situ Monitoring of Fatigue Crack Progression

    NASA Astrophysics Data System (ADS)

    Michaels, J. E.; Michaels, T. E.; Mi, B.; Cobb, A. C.; Stobbe, D. M.

    2005-04-01

    Ultrasonic sensors permanently affixed to aluminum coupons are used to monitor progression of damage during fatigue testing with the long term goal of structural health monitoring for diagnostics and prognostics. Necessary for success are proper design of the ultrasonic testing methods, robust transducer mounting techniques, and real-time signal processing for determining the state of the structure. It is also highly desirable for the overall system to be self-calibrating with built-in diagnostics in order to detect and compensate for sensor degradation or failure. Self-calibrating ultrasonic techniques are applied for monitoring of cracks initiating and propagating from the inaccessible inner diameters of rivet holes where the transducers are mounted on the accessible specimen surface. Angle beam ultrasonic methods are utilized that are suitable for detecting small defects in critical local regions of high stress. Results are presented for aluminum coupons subjected to low cycle fatigue and demonstrate ultrasonic tracking of crack growth.

  16. Calibration method and apparatus for measuring the concentration of components in a fluid

    DOEpatents

    Durham, Michael D.; Sagan, Francis J.; Burkhardt, Mark R.

    1993-01-01

    A calibration method and apparatus for use in measuring the concentrations of components of a fluid is provided. The measurements are determined from the intensity of radiation over a selected range of radiation wavelengths using peak-to-trough calculations. The peak-to-trough calculations are simplified by compensating for radiation absorption by the apparatus. The invention also allows absorption characteristics of an interfering fluid component to be accurately determined and negated thereby facilitating analysis of the fluid.

  17. Calibration method and apparatus for measuring the concentration of components in a fluid

    DOEpatents

    Durham, M.D.; Sagan, F.J.; Burkhardt, M.R.

    1993-12-21

    A calibration method and apparatus for use in measuring the concentrations of components of a fluid is provided. The measurements are determined from the intensity of radiation over a selected range of radiation wavelengths using peak-to-trough calculations. The peak-to-trough calculations are simplified by compensating for radiation absorption by the apparatus. The invention also allows absorption characteristics of an interfering fluid component to be accurately determined and negated thereby facilitating analysis of the fluid. 7 figures.

  18. Solid form additives and method of forming same

    SciTech Connect

    Schuettenberg, A.D.; Gragson, J.T.

    1987-01-27

    This patent describes a solid form additive comprising: a normally liquid fuel additive selected from carburetor detergent additives, antiknock additives, deposit-control additives, and mixtures thereof, suitable for use in fuel comprising gasoline for internal combustion engines; and a structural agent for containing the fuel additive and for providing dimensional stability to the solid form additive, the structural agent being soluble and dispersible in the fuel; wherein the fuel additive comprises between about 25% and about 75% by weight of the solid form additive; and wherein the solid form additive is a pellet having structural agent and fuel additive essentially homogeneously dispersed throughout the solid form additive; and wherein the pellet is coated with a sealing agent.

  19. Application of Response Surface based Calibration and Sensitivity Analysis methods for Regional Hydrogeological Modelling in the Western Canada Sedimentary Basin

    NASA Astrophysics Data System (ADS)

    Singh, A.; Palombi, D.; Huff, G. F.

    2014-12-01

    A regional scale study of groundwater flow dynamics was undertaken in the Western Canada Sedimentary Basin (WCSB), comprising parts of Alberta, Saskatchewan and British Columbia. The objective of the study is to investigate basin-scale hydrogeology in WCSB and to establish boundary conditions for future local-scale groundwater management models. Earlier work in the Alberta basin has acknowledged the fact that in addition to topography controlled conditions, a substantial part of the basin exhibits sub-hydrostatic regimes. The basin-scale model (approx. 420,000 km2) includes Upper Cretaceous aquifers to Recent age sediments which collectively attain maximum thicknesses of >2600 m. Regional aquifer units considered for the numerical model are Quaternary sediments, and the sedimentary rocks of the Paskapoo, Scollard, Horseshoe Canyon formations and the Belly River Group. Regional aquitards delineated include the Battle and Bear Paw formations. The study area is bound to the west by the Brazeau-Waptiti thrust belt and to the south by the Canada-USA international border. The boundary to the north and east is delineated by the maximum extent of the Wapiti and Belly River groups and Judith River Formation. USGS MODFLOW was implemented for numerical simulation. The steady state numerical model was calibrated using a Response Surface based (Radial Basis Functions) optimization method. The calibration targets (~2000) were comprised of drill stem tests for deeper units and static water levels for shallower units. Petrophysical analyses of cores averaged K values from analyses of aquifer test results,and literature values were used to provide initial values and calibration ranges for hydraulic properties. Results indicate predominance of topography driven, local- to intermediate-scale flow systems in all hydrostratigraphic units with recharge of these units occurring in the foothills of the Rocky Mountains. The Battle aquitard, where present, acts to retard regional flow

  20. A Vision-Based Self-Calibration Method for Robotic Visual Inspection Systems

    PubMed Central

    Yin, Shibin; Ren, Yongjie; Zhu, Jigui; Yang, Shourui; Ye, Shenghua

    2013-01-01

    A vision-based robot self-calibration method is proposed in this paper to evaluate the kinematic parameter errors of a robot using a visual sensor mounted on its end-effector. This approach could be performed in the industrial field without external, expensive apparatus or an elaborate setup. A robot Tool Center Point (TCP) is defined in the structural model of a line-structured laser sensor, and aligned to a reference point fixed in the robot workspace. A mathematical model is established to formulate the misalignment errors with kinematic parameter errors and TCP position errors. Based on the fixed point constraints, the kinematic parameter errors and TCP position errors are identified with an iterative algorithm. Compared to the conventional methods, this proposed method eliminates the need for a robot-based-frame and hand-to-eye calibrations, shortens the error propagation chain, and makes the calibration process more accurate and convenient. A validation experiment is performed on an ABB IRB2400 robot. An optimal configuration on the number and distribution of fixed points in the robot workspace is obtained based on the experimental results. Comparative experiments reveal that there is a significant improvement of the measuring accuracy of the robotic visual inspection system. PMID:24300597

  1. Atomic Force Microscope Cantilever Flexural Stiffness Calibration: Toward a Standard Traceable Method

    PubMed Central

    Gates, Richard S.; Reitsma, Mark G.; Kramar, John A.; Pratt, Jon R.

    2011-01-01

    The evolution of the atomic force microscope into a useful tool for measuring mechanical properties of surfaces at the nanoscale has spurred the need for more precise and accurate methods for calibrating the spring constants of test cantilevers. Groups within international standards organizations such as the International Organization for Standardization and the Versailles Project on Advanced Materials and Standards (VAMAS) are conducting studies to determine which methods are best suited for these calibrations and to try to improve the reproducibility and accuracy of these measurements among different laboratories. This paper expands on a recent mini round robin within VAMAS Technical Working Area 29 to measure the spring constant of a single batch of triangular silicon nitride cantilevers sent to three international collaborators. Calibration techniques included reference cantilever, added mass, and two forms of thermal methods. Results are compared to measurements traceable to the International System of Units provided by an electrostatic force balance. A series of guidelines are also discussed for procedures that can improve the running of round robins in atomic force microscopy. PMID:26989594

  2. A novel method of strain - bending moment calibration for blade testing

    NASA Astrophysics Data System (ADS)

    Greaves, P.; Prieto, R.; Gaffing, J.; van Beveren, C.; Dominy, R.; Ingram, G.

    2016-09-01

    A new method of interpreting strain data in full scale static and fatigue tests has been implemented as part of the Offshore Renewable Energy Catapult's ongoing development of biaxial fatigue testing of wind turbine blades. During bi-axial fatigue tests, it is necessary to be able to distinguish strains arising from the flapwise motion of the blade from strains arising from the edgewise motion. The method exploits the beam-like structure of blades and is derived using the equations of beam theory. It offers several advantages over the current state of the art method of calibrating strain gauges.

  3. Radiometric calibration of SPOT 2 HRV - A comparison of three methods

    NASA Technical Reports Server (NTRS)

    Biggar, Stuart F.; Dinguirard, Magdeleine C.; Gellman, David I.; Henry, Patrice; Jackson, Ray D.; Moran, M. S.; Slater, Philip N.

    1991-01-01

    Three methods for determining an absolute radiometric calibration of a spacecraft optical sensor are compared. They are the well-known reflectance-based and radiance-based methods and a new method based on measurements of the ratio of diffuse-to-global irradiance at the ground. The latter will be described in detail and the comparison of the three approaches will be made with reference to the SPOT-2 HRV cameras for a field campaign 1990-06-19 through 1990-06-24 at the White Sands Missile Range in New Mexico.

  4. Accurate and automatic extrinsic calibration method for blade measurement system integrated by different optical sensors

    NASA Astrophysics Data System (ADS)

    He, Wantao; Li, Zhongwei; Zhong, Kai; Shi, Yusheng; Zhao, Can; Cheng, Xu

    2014-11-01

    Fast and precise 3D inspection system is in great demand in modern manufacturing processes. At present, the available sensors have their own pros and cons, and hardly exist an omnipotent sensor to handle the complex inspection task in an accurate and effective way. The prevailing solution is integrating multiple sensors and taking advantages of their strengths. For obtaining a holistic 3D profile, the data from different sensors should be registrated into a coherent coordinate system. However, some complex shape objects own thin wall feather such as blades, the ICP registration method would become unstable. Therefore, it is very important to calibrate the extrinsic parameters of each sensor in the integrated measurement system. This paper proposed an accurate and automatic extrinsic parameter calibration method for blade measurement system integrated by different optical sensors. In this system, fringe projection sensor (FPS) and conoscopic holography sensor (CHS) is integrated into a multi-axis motion platform, and the sensors can be optimally move to any desired position at the object's surface. In order to simple the calibration process, a special calibration artifact is designed according to the characteristics of the two sensors. An automatic registration procedure based on correlation and segmentation is used to realize the artifact datasets obtaining by FPS and CHS rough alignment without any manual operation and data pro-processing, and then the Generalized Gauss-Markoff model is used to estimate the optimization transformation parameters. The experiments show the measurement result of a blade, where several sampled patches are merged into one point cloud, and it verifies the performance of the proposed method.

  5. Calibration method of the time synchronization error of many data acquisition nodes in the chained system

    NASA Astrophysics Data System (ADS)

    Jiang, Jia-jia; Duan, Fa-jie; Chen, Jin; Zhang, Chao; Wang, Kai; Chang, Zong-jie

    2012-08-01

    Time synchronization is very important in a distributed chained seismic acquisition system with a large number of data acquisition nodes (DANs). The time synchronization error has two causes. On the one hand, there is a large accumulated propagation delay when commands propagate from the analysis and control system to multiple distant DANs, which makes it impossible for different DANs to receive the same command synchronously. Unfortunately, the propagation delay of commands (PDCs) varies in different application environments. On the other hand, the phase jitter of both the master clock and the clock recovery phase-locked loop, which is designed to extract the timing signal, may also cause the time synchronization error. In this paper, in order to achieve accurate time synchronization, a novel calibration method is proposed which can align the PDCs of all of the DANs in real time and overcome the time synchronization error caused by the phase jitter. Firstly, we give a quantitative analysis of the time synchronization error caused by both the PDCs and the phase jitter. Secondly, we propose a back and forth model (BFM) and a transmission delay measurement method (TDMM) to overcome these difficulties. Furthermore, the BFM is designed as the hardware configuration to measure the PDCs and calibrate the time synchronization error. The TDMM is used to measure the PDCs accurately. Thirdly, in order to overcome the time synchronization error caused by the phase jitter, a compression and mapping algorithm (CMA) is presented. Finally, based on the proposed BFM, TDMM and CMA, a united calibration algorithm is developed to overcome the time synchronization error caused by both the PDCs and the phase jitter. The simulation experiment results show the effectiveness of the calibration method proposed in this paper.

  6. Nonlinearity measurements of solar cells with an LED-based combinatorial flux addition method

    PubMed Central

    Hamadani, Behrang H.; Shore, Andrew; Roller, John; Yoon, Howard W; Campanelli, Mark

    2016-01-01

    We present a light emitting diode (LED)-based system utilizing a combinatorial flux addition method to investigate the nonlinear relationship in solar cells between the output current of the cell and the incident irradiance level. The magnitude of the light flux is controlled by the supplied currents to two LEDs (or two sets of them) in a combinatorial fashion. The signals measured from the cell are arranged within a related overdetermined linear system of equations derived from an appropriately chosen Nth degree polynomial representing the relationship between the measured signals and the incident fluxes. The flux values and the polynomial coefficients are then solved for by linear least squares to obtain the best fit. The technique can be applied to any solar cell, under either monochromatic or broadband spectrum. For the unscaled solution, no reference detectors or prior calibrations of the light flux are required. However, if at least one calibrated irradiance value is known, then the entire curve can be scaled to an appropriate spectral responsivity value. Using this technique, a large number of data points can be obtained in a relatively short time scale over a large signal range. PMID:27524837

  7. Nonlinearity measurements of solar cells with an LED-based combinatorial flux addition method.

    PubMed

    Hamadani, Behrang H; Shore, Andrew; Roller, John; Yoon, Howard W; Campanelli, Mark

    2016-02-01

    We present a light emitting diode (LED)-based system utilizing a combinatorial flux addition method to investigate the nonlinear relationship in solar cells between the output current of the cell and the incident irradiance level. The magnitude of the light flux is controlled by the supplied currents to two LEDs (or two sets of them) in a combinatorial fashion. The signals measured from the cell are arranged within a related overdetermined linear system of equations derived from an appropriately chosen N(th) degree polynomial representing the relationship between the measured signals and the incident fluxes. The flux values and the polynomial coefficients are then solved for by linear least squares to obtain the best fit. The technique can be applied to any solar cell, under either monochromatic or broadband spectrum. For the unscaled solution, no reference detectors or prior calibrations of the light flux are required. However, if at least one calibrated irradiance value is known, then the entire curve can be scaled to an appropriate spectral responsivity value. Using this technique, a large number of data points can be obtained in a relatively short time scale over a large signal range.

  8. Calibration method for synchronizing the viewing zones of a large-scale autostereoscopic multi-view laser display

    NASA Astrophysics Data System (ADS)

    Reitterer, Jörg; Fidler, Franz; Schmid, Gerhard; Hambeck, Christian; Saint Julien-Wallsee, Ferdinand; Leeb, Walter; Schmid, Ulrich

    2015-06-01

    We have developed a calibration method for synchronizing the autostereoscopic viewing zones displayed by an array of "trixels", i.e., laser modules with integrated MEMS mirrors. Calibration patterns are projected onto a calibration screen and recorded by a camera. The distortion functions, which map the laser diode driving signals from the time domain to the intensity distributions in the spatial domain, are extracted from the recorded calibration patterns. Unlike conventional autostereoscopic displays, e.g., with lenticular lenses or parallax barriers, the optimal 3D viewing distance of the display can be adapted on-the-fly by transforming the distortion functions using linear operations.

  9. Flight Test Results of a GPS-Based Pitot-Static Calibration Method Using Output-Error Optimization for a Light Twin-Engine Airplane

    NASA Technical Reports Server (NTRS)

    Martos, Borja; Kiszely, Paul; Foster, John V.

    2011-01-01

    As part of the NASA Aviation Safety Program (AvSP), a novel pitot-static calibration method was developed to allow rapid in-flight calibration for subscale aircraft while flying within confined test areas. This approach uses Global Positioning System (GPS) technology coupled with modern system identification methods that rapidly computes optimal pressure error models over a range of airspeed with defined confidence bounds. This method has been demonstrated in subscale flight tests and has shown small 2- error bounds with significant reduction in test time compared to other methods. The current research was motivated by the desire to further evaluate and develop this method for full-scale aircraft. A goal of this research was to develop an accurate calibration method that enables reductions in test equipment and flight time, thus reducing costs. The approach involved analysis of data acquisition requirements, development of efficient flight patterns, and analysis of pressure error models based on system identification methods. Flight tests were conducted at The University of Tennessee Space Institute (UTSI) utilizing an instrumented Piper Navajo research aircraft. In addition, the UTSI engineering flight simulator was used to investigate test maneuver requirements and handling qualities issues associated with this technique. This paper provides a summary of piloted simulation and flight test results that illustrates the performance and capabilities of the NASA calibration method. Discussion of maneuver requirements and data analysis methods is included as well as recommendations for piloting technique.

  10. Fuzzy Filtering Method for Color Videos Corrupted by Additive Noise

    PubMed Central

    Ponomaryov, Volodymyr I.; Montenegro-Monroy, Hector; Nino-de-Rivera, Luis

    2014-01-01

    A novel method for the denoising of color videos corrupted by additive noise is presented in this paper. The proposed technique consists of three principal filtering steps: spatial, spatiotemporal, and spatial postprocessing. In contrast to other state-of-the-art algorithms, during the first spatial step, the eight gradient values in different directions for pixels located in the vicinity of a central pixel as well as the R, G, and B channel correlation between the analogous pixels in different color bands are taken into account. These gradient values give the information about the level of contamination then the designed fuzzy rules are used to preserve the image features (textures, edges, sharpness, chromatic properties, etc.). In the second step, two neighboring video frames are processed together. Possible local motions between neighboring frames are estimated using block matching procedure in eight directions to perform interframe filtering. In the final step, the edges and smoothed regions in a current frame are distinguished for final postprocessing filtering. Numerous simulation results confirm that this novel 3D fuzzy method performs better than other state-of-the-art techniques in terms of objective criteria (PSNR, MAE, NCD, and SSIM) as well as subjective perception via the human vision system in the different color videos. PMID:24688428

  11. A method for small-animal PET/CT alignment calibration

    NASA Astrophysics Data System (ADS)

    Pascau, J.; Vaquero, J. J.; Chamorro-Servent, J.; Rodríguez-Ruano, A.; Desco, M.

    2012-06-01

    Small-animal positron-emission tomography/computed tomography (PET/CT) scanners provide anatomical and molecular imaging, which enables the joint visualization and analysis of both types of data. A proper alignment calibration procedure is essential for small-animal imaging since resolution is much higher than that in human devices. This work presents an alignment phantom and two different calibration methods that provide a reliable and repeatable measurement of the spatial geometrical alignment between the PET and the CT subsystems of a hybrid scanner. The phantom can be built using laboratory materials, and it is meant to estimate the rigid spatial transformation that aligns both modalities. It consists of three glass capillaries filled with a positron-emitter solution and positioned in a non-coplanar triangular geometry inside the system field of view. The calibration methods proposed are both based on automatic line detection, but with different approaches to calculate the transformation of the lines between both modalities. Our results show an average accuracy of the alignment estimation of 0.39 mm over the whole field of view.

  12. In-Flight Calibration Methods for Temperature-Dependent Offsets in the MMS Fluxgate Magnetometers

    NASA Technical Reports Server (NTRS)

    Bromund, K. R.; Plaschke, F.; Strangeway, R. J.; Anderson, B. J.; Huang, B. G.; Magnes, W.; Fischer, D.; Nakamura, R.; Leinweber, H. K.; Russell, C. T.; Baumjohann, W.; Chutter, M.; Torbert, R. B.; Le, G.; Slavin, J. A.; Kepko, E. L.

    2016-01-01

    During the first dayside season of the Magnetospheric Multiscale (MMS) mission, the in-flight calibration process for the Fluxgate magnetometers (FGM) implemented an algorithm that selected a constant offset (zero-level) for each sensor on each orbit. This method was generally able to reduce the amplitude of residual spin tone to less than 0.2 nT within the region of interest. However, there are times when the offsets do show significant short-term variations. These variations are most prominent in the nighttime season (phase 1X), when eclipses are accompanied by offset changes as large as 1 nT. Eclipses are followed by a recovery period as long as 12 hours where the offsets continue to change as temperatures stabilize. Understanding and compensating for these changes will become critical during Phase 2 of the mission in 2017, when the nightside will become the focus of MMS science. Although there is no direct correlation between offset and temperature, the offsets are seen for the period of any given week to be well-characterized as function of instrument temperature. Using this property, a new calibration method has been developed that has proven effective in compensating for temperature-dependent offsets during phase 1X of the MMS mission and also promises to further refine calibration quality during the dayside season.

  13. Dew Point Calibration System Using a Quartz Crystal Sensor with a Differential Frequency Method

    PubMed Central

    Lin, Ningning; Meng, Xiaofeng; Nie, Jing

    2016-01-01

    In this paper, the influence of temperature on quartz crystal microbalance (QCM) sensor response during dew point calibration is investigated. The aim is to present a compensation method to eliminate temperature impact on frequency acquisition. A new sensitive structure is proposed with double QCMs. One is kept in contact with the environment, whereas the other is not exposed to the atmosphere. There is a thermal conductivity silicone pad between each crystal and a refrigeration device to keep a uniform temperature condition. A differential frequency method is described in detail and is applied to calibrate the frequency characteristics of QCM at the dew point of −3.75 °C. It is worth noting that frequency changes of two QCMs were approximately opposite when temperature conditions were changed simultaneously. The results from continuous experiments show that the frequencies of two QCMs as the dew point moment was reached have strong consistency and high repeatability, leading to the conclusion that the sensitive structure can calibrate dew points with high reliability. PMID:27869746

  14. Research on the method of establishing the total radiation meter calibration device

    NASA Astrophysics Data System (ADS)

    Gao, Jianqiang; Xia, Ming; Xia, Junwen; Zhang, Dong

    2015-10-01

    Pyranometer is an instrument used to measure the solar radiation, according to pyranometer differs as installation state, can be respectively measured total solar radiation, reflected radiation, or with the help of shading device for measuring scattering radiation. Pyranometer uses the principle of thermoelectric effect, inductive element adopts winding plating type multi junction thermopile, its surface is coated with black coating with high absorption rate. Hot junction in the induction surface, while the cold junction is located in the body, the cold and hot junction produce thermoelectric potential. In the linear range, the output signal is proportional to the solar irradiance. Traceability to national meteorological station, as the unit of the national legal metrology organizations, the responsibility is to transfer value of the sun and the earth radiation value about the national meteorological industry. Using the method of comparison, with indoor calibration of solar simulator, at the same location, standard pyranometer and measured pyranometer were alternately measured radiation irradiance, depending on the irradiation sensitivity standard pyranometer were calculated the radiation sensitivity of measured pyranometer. This paper is mainly about the design and calibration method of the pyranometer indoor device. The uncertainty of the calibration result is also evaluated.

  15. Methods for detecting additional genes underlying Alzheimer disease

    SciTech Connect

    Locke, P.A.; Haines, J.L.; Ter-Minassian, M.

    1994-09-01

    Alzheimer`s disease (AD) is a complex inherited disorder with proven genetic heterogeneity. To date, genes on chromosome 21 (APP) and 14 (not yet identified) are associated with early-onset familial AD, while the APOE gene on chromosome 19 is associated with both late onset familial and sporadic AD and early onset sporadic AD. Although these genes likely account for the majority of AD, many familial cases cannot be traced to any of these genes. From a set of 127 late-onset multiplex families screened for APOE, 43 (34%) families have at least one affected individual with no APOE-4 allele, suggesting an alternative genetic etiology. Simulation studies indicated that additional loci could be identified through a genomic screen with a 10 cM sieve on a subset of 21 well documented, non-APOE-4 families. Given the uncertainties in the mode of inheritance, reliance on a single analytical method could result in a missed linkage. Therefore, we have developed a strategy of using multiple overlapping yet complementary methods to detect linkage. These include sib-pair analysis and affected-pedigree-member analysis, neither of which makes assumptions about mode of inheritance, and lod score analysis (using two predefined genetic models). In order for a marker to qualify for follow-up, it must fit at least two of three criteria. These are nominal P values of 0.05 or less for the non-parametric methods, and/or a lod score greater than 1.0. Adjacent markers each fulfilling a single criterion also warrant follow-up. To date, we have screened 61 markers on chromosomes 1, 2, 3, 18, 19, 21, and 22. One marker, D2S163, generated a lod score of 1.06 ({theta} = 0.15) and an APMT statistic of 3.68 (P < 0.001). This region is currently being investigated in more detail. Updated results of this region plus additional screening data will be presented.

  16. High accuracy position response calibration method for a micro-channel plate ion detector

    NASA Astrophysics Data System (ADS)

    Hong, R.; Leredde, A.; Bagdasarova, Y.; Fléchard, X.; García, A.; Müller, P.; Knecht, A.; Liénard, E.; Kossin, M.; Sternberg, M. G.; Swanson, H. E.; Zumwalt, D. W.

    2016-11-01

    We have developed a position response calibration method for a micro-channel plate (MCP) detector with a delay-line anode position readout scheme. Using an in situ calibration mask, an accuracy of 8 μm and a resolution of 85 μm (FWHM) have been achieved for MeV-scale α particles and ions with energies of ∼10 keV. At this level of accuracy, the difference between the MCP position responses to high-energy α particles and low-energy ions is significant. The improved performance of the MCP detector can find applications in many fields of AMO and nuclear physics. In our case, it helps reducing systematic uncertainties in a high-precision nuclear β-decay experiment.

  17. Neutron monitoring systems including gamma thermometers and methods of calibrating nuclear instruments using gamma thermometers

    DOEpatents

    Moen, Stephan Craig; Meyers, Craig Glenn; Petzen, John Alexander; Foard, Adam Muhling

    2012-08-07

    A method of calibrating a nuclear instrument using a gamma thermometer may include: measuring, in the instrument, local neutron flux; generating, from the instrument, a first signal proportional to the neutron flux; measuring, in the gamma thermometer, local gamma flux; generating, from the gamma thermometer, a second signal proportional to the gamma flux; compensating the second signal; and calibrating a gain of the instrument based on the compensated second signal. Compensating the second signal may include: calculating selected yield fractions for specific groups of delayed gamma sources; calculating time constants for the specific groups; calculating a third signal that corresponds to delayed local gamma flux based on the selected yield fractions and time constants; and calculating the compensated second signal by subtracting the third signal from the second signal. The specific groups may have decay time constants greater than 5.times.10.sup.-1 seconds and less than 5.times.10.sup.5 seconds.

  18. Note: A calibration method to determine the lumped-circuit parameters of a magnetic probe.

    PubMed

    Li, Fuming; Chen, Zhipeng; Zhu, Lizhi; Liu, Hai; Wang, Zhijiang; Zhuang, Ge

    2016-06-01

    This paper describes a novel method to determine the lumped-circuit parameters of a magnetic inductive probe for calibration by using Helmholtz coils with high frequency power supply (frequency range: 10 kHz-400 kHz). The whole calibration circuit system can be separated into two parts: "generator" circuit and "receiver" circuit. By implementing the Fourier transform, two analytical lumped-circuit models, with respect to these separated circuits, are constructed to obtain the transfer function between each other. Herein, the precise lumped-circuit parameters (including the resistance, inductance, and capacitance) of the magnetic probe can be determined by fitting the experimental data to the transfer function. Regarding the fitting results, the finite impedance of magnetic probe can be used to analyze the transmission of a high-frequency signal between magnetic probes, cables, and acquisition system.

  19. Note: A calibration method to determine the lumped-circuit parameters of a magnetic probe

    NASA Astrophysics Data System (ADS)

    Li, Fuming; Chen, Zhipeng; Zhu, Lizhi; Liu, Hai; Wang, Zhijiang; Zhuang, Ge

    2016-06-01

    This paper describes a novel method to determine the lumped-circuit parameters of a magnetic inductive probe for calibration by using Helmholtz coils with high frequency power supply (frequency range: 10 kHz-400 kHz). The whole calibration circuit system can be separated into two parts: "generator" circuit and "receiver" circuit. By implementing the Fourier transform, two analytical lumped-circuit models, with respect to these separated circuits, are constructed to obtain the transfer function between each other. Herein, the precise lumped-circuit parameters (including the resistance, inductance, and capacitance) of the magnetic probe can be determined by fitting the experimental data to the transfer function. Regarding the fitting results, the finite impedance of magnetic probe can be used to analyze the transmission of a high-frequency signal between magnetic probes, cables, and acquisition system.

  20. The development of in-situ calibration method for divertor IR thermography in ITER

    SciTech Connect

    Takeuchi, M.; Sugie, T.; Ogawa, H.; Takeyama, S.; Itami, K.

    2014-08-21

    For the development of the calibration method of the emissivity in IR light on the divertor plate in ITER divertor IR thermography system, the laboratory experiments have been performed by using IR instruments. The calibration of the IR camera was performed by the plane black body in the temperature of 100–600 degC. The radiances of the tungsten heated by 280 degC were measured by the IR camera without filter (2.5–5.1 μm) and with filter (2.95 μm, 4.67 μm). The preliminary data of the scattered light of the laser of 3.34 μm that injected into the tungsten were acquired.

  1. Heterodyne interferometry method for calibration of a Soleil-Babinet compensator.

    PubMed

    Zhang, Wenjing; Zhang, Zhiwei

    2016-05-20

    A method based on the common-path heterodyne interferometer system is proposed for the calibration of a Soleil-Babinet compensator. In this heterodyne interferometer system, which consists of two acousto-optic modulators, the compensator being calibrated is inserted into the signal path. By using the reference beam as the benchmark and a lock-in amplifier (SR844) as the phase retardation collector, retardations of 0 and λ (one wavelength) can be located accurately, and an arbitrary retardation between 0 and λ can also be measured accurately and continuously. By fitting a straight line to the experimental data, we obtained a linear correlation coefficient (R) of 0.995, which indicates that this system is capable of linear phase detection. The experimental results demonstrate determination accuracies of 0.212° and 0.26° and measurement precisions of 0.054° and 0.608° for retardations of 0 and λ, respectively.

  2. Easy and direct method for calibrating atomic force microscopy lateral force measurements

    PubMed Central

    Liu, Wenhua; Bonin, Keith; Guthold, Martin

    2010-01-01

    We have designed and tested a new, inexpensive, easy-to-make and easy-to-use calibration standard for atomic force microscopy (AFM) lateral force measurements. This new standard simply consists of a small glass fiber of known dimensions and Young’s modulus, which is fixed at one end to a substrate and which can be bent laterally with the AFM tip at the other end. This standard has equal or less error than the commonly used method of using beam mechanics to determine a cantilever’s lateral force constant. It is transferable, thus providing a universal tool for comparing the calibrations of different instruments. It does not require knowledge of the cantilever dimensions and composition or its tip height. This standard also allows direct conversion of the photodiode signal to force and, thus, circumvents the requirement for a sensor response (sensitivity) measurement. PMID:17614616

  3. On-orbit calibration approach for star cameras based on the iteration method with variable weights.

    PubMed

    Wang, Mi; Cheng, Yufeng; Yang, Bo; Chen, Xiao

    2015-07-20

    To perform efficient on-orbit calibration for star cameras, we developed an attitude-independent calibration approach for global optimization and noise removal by least-square estimation using multiple star images, with which the optimal principal point, focal length, and the high-order focal plane distortion can be obtained in one step in full consideration of the interaction among star camera parameters. To avoid the problem when stars could be misidentified in star images, an iteration method with variable weights is introduced to eliminate the influence of misidentified star pairs. The approach can increase the precision of least-square estimation and use fewer star images. The proposed approach has been well verified to be precise and robust in three experiments.

  4. A method for automating calibration and records management for instrumentation and dosimetry

    SciTech Connect

    O`Brien, J.M. Jr.; Rushton, R.O.; Burns, R.E. Jr.

    1993-12-31

    Current industry requirements are becoming more stringent on quality assurance records and documentation for calibration of instruments and dosimetry. A novel method is presented here that will allow a progressive automation scheme to be used in pursuit of that goal. This concept is based on computer-controlled irradiators that can act as stand-alone devices or be interfaced to other components via a computer local area network. In this way, complete systems can be built with modules to create a records management system to meet the needs of small laboratories or large multi-building calibration groups. Different database engines or formats can be used simply by replacing a module. Modules for temperature and pressure monitoring or shipping and receiving can be added, as well as equipment modules for direct IEEE-488 interface to electrometers and other instrumentation.

  5. [The use of different albumin preparations as calibrators in determining the total protein in blood serum by the biuret method].

    PubMed

    Sigalov, A B; Isaeva, N V; Bezruchkina, S V

    1993-01-01

    The authors have investigated the possibility of using various albumin preparations as calibrators in measurements of human blood serum total protein by the biuret method. Analysis of Precinorm U and Precipath U reference sera has demonstrated that use of various albumin preparations as calibrators may result in significant deviations (as much as 27%) of the resultant values from the due ones.

  6. A calibration method for lateral forces for use with colloidal probe force microscopy cantilevers

    SciTech Connect

    Quintanilla, M. A. S.; Goddard, D. T.

    2008-02-15

    A calibration method is described for colloidal probe cantilevers that enables friction force measurements obtained using lateral force microscopy (LFM) to be quantified. The method is an adaptation of the lever method of Feiler et al. [A. Feiler, P. Attard, and I. Larson, Rev. Sci. Instum. 71, 2746 (2000)] and uses the advantageous positioning of probe particles that are usually offset from the central axis of the cantilever. The main sources of error in the calibration method are assessed, in particular, the potential misalignment of the long axis of the cantilever that ideally should be perpendicular to the photodiode detector. When this is not taken into account, the misalignment is shown to have a significant effect on the cantilever torsional stiffness but not on the lateral photodiode sensitivity. Also, because the friction signal is affected by the topography of the substrate, the method presented is valid only against flat substrates. Two types of particles, 20 {mu}m glass beads and UO{sub 3} agglomerates attached to silicon tapping mode cantilevers were used to test the method against substrates including glass, cleaved mica, and UO{sub 2} single crystals. Comparisons with the lateral compliance method of Cain et al. [R. G. Cain, S. Biggs, and N. W. Page, J. Colloid Interface Sci. 227, 55 (2000)] are also made.

  7. Considerations on the quinine actinometry calibration method used in photostability testing of pharmaceuticals.

    PubMed

    de Azevedo Filho, Clayton Anderson; de Filgueiras Gomes, Daniel; de Mélo Guedes, João Paulo; Batista, Rayana Muniz Falcão; Santos, Beate Saegesser

    2011-03-25

    This paper addresses two critical issues concerning the guidelines adopted by the ICH on the photostability testing: the quinine actinometry method and the light/radiation exposure map distribution of the photostability chamber. Using a qualified non-commercial photostability chamber tests were performed using quinine and physical actinometry and compared the results to those which are used as the basis of the ICH guidelines. The statistical analysis on the results showed that: (i) the calibration curve of the quinine solution depends on its concentration and on its location in the chamber; (ii) the quinine actinometry method currently recommended by the ICH guidelines should not be generalized to any photostability chamber.

  8. A method for the comparison of performance of gamma-ray spectrometry calibration cocktails.

    PubMed

    Legarda, F; Los Arcos, J M; Herranz, M

    2004-01-01

    In order to make quantitative assessments about the usefulness of different gamma-ray emitting radionuclide cocktails to carry out efficiency calibrations of gamma-ray spectrometers, a method has been developed that allows the comparison of their different performances and to optimize the choice of gamma energy lines for the radionuclides within a specific cocktail. The method has been applied to compare different cocktail configurations obtained from measurements made in the laboratory with monoenergetic radionuclides, and their relative performances are presented and discussed.

  9. A Rapid Coordinate Transformation Method Applied in Industrial Robot Calibration Based on Characteristic Line Coincidence.

    PubMed

    Liu, Bailing; Zhang, Fumin; Qu, Xinghua; Shi, Xiaojia

    2016-02-18

    Coordinate transformation plays an indispensable role in industrial measurements, including photogrammetry, geodesy, laser 3-D measurement and robotics. The widely applied methods of coordinate transformation are generally based on solving the equations of point clouds. Despite the high accuracy, this might result in no solution due to the use of ill conditioned matrices. In this paper, a novel coordinate transformation method is proposed, not based on the equation solution but based on the geometric transformation. We construct characteristic lines to represent the coordinate systems. According to the space geometry relation, the characteristic line scan is made to coincide by a series of rotations and translations. The transformation matrix can be obtained using matrix transformation theory. Experiments are designed to compare the proposed method with other methods. The results show that the proposed method has the same high accuracy, but the operation is more convenient and flexible. A multi-sensor combined measurement system is also presented to improve the position accuracy of a robot with the calibration of the robot kinematic parameters. Experimental verification shows that the position accuracy of robot manipulator is improved by 45.8% with the proposed method and robot calibration.

  10. A Novel Calibration-Minimum Method for Prediction of Mole Fraction in Non-Ideal Mixture.

    PubMed

    Shibayama, Shojiro; Kaneko, Hiromasa; Funatsu, Kimito

    2017-04-01

    This article proposes a novel concentration prediction model that requires little training data and is useful for rapid process understanding. Process analytical technology is currently popular, especially in the pharmaceutical industry, for enhancement of process understanding and process control. A calibration-free method, iterative optimization technology (IOT), was proposed to predict pure component concentrations, because calibration methods such as partial least squares, require a large number of training samples, leading to high costs. However, IOT cannot be applied to concentration prediction in non-ideal mixtures because its basic equation is derived from the Beer-Lambert law, which cannot be applied to non-ideal mixtures. We proposed a novel method that realizes prediction of pure component concentrations in mixtures from a small number of training samples, assuming that spectral changes arising from molecular interactions can be expressed as a function of concentration. The proposed method is named IOT with virtual molecular interaction spectra (IOT-VIS) because the method takes spectral change as a virtual spectrum x nonlin,i into account. It was confirmed through the two case studies that the predictive accuracy of IOT-VIS was the highest among existing IOT methods.

  11. A Rapid Coordinate Transformation Method Applied in Industrial Robot Calibration Based on Characteristic Line Coincidence

    PubMed Central

    Liu, Bailing; Zhang, Fumin; Qu, Xinghua; Shi, Xiaojia

    2016-01-01

    Coordinate transformation plays an indispensable role in industrial measurements, including photogrammetry, geodesy, laser 3-D measurement and robotics. The widely applied methods of coordinate transformation are generally based on solving the equations of point clouds. Despite the high accuracy, this might result in no solution due to the use of ill conditioned matrices. In this paper, a novel coordinate transformation method is proposed, not based on the equation solution but based on the geometric transformation. We construct characteristic lines to represent the coordinate systems. According to the space geometry relation, the characteristic line scan is made to coincide by a series of rotations and translations. The transformation matrix can be obtained using matrix transformation theory. Experiments are designed to compare the proposed method with other methods. The results show that the proposed method has the same high accuracy, but the operation is more convenient and flexible. A multi-sensor combined measurement system is also presented to improve the position accuracy of a robot with the calibration of the robot kinematic parameters. Experimental verification shows that the position accuracy of robot manipulator is improved by 45.8% with the proposed method and robot calibration. PMID:26901203

  12. An efficient surrogate-based simulation-optimization method for calibrating a regional MODFLOW model

    NASA Astrophysics Data System (ADS)

    Chen, Mingjie; Izady, Azizallah; Abdalla, Osman A.

    2017-01-01

    Simulation-optimization method entails a large number of model simulations, which is computationally intensive or even prohibitive if the model simulation is extremely time-consuming. Statistical models have been examined as a surrogate of the high-fidelity physical model during simulation-optimization process to tackle this problem. Among them, Multivariate Adaptive Regression Splines (MARS), a non-parametric adaptive regression method, is superior in overcoming problems of high-dimensions and discontinuities of the data. Furthermore, the stability and accuracy of MARS model can be improved by bootstrap aggregating methods, namely, bagging. In this paper, Bagging MARS (BMARS) method is integrated to a surrogate-based simulation-optimization framework to calibrate a three-dimensional MODFLOW model, which is developed to simulate the groundwater flow in an arid hardrock-alluvium region in northwestern Oman. The physical MODFLOW model is surrogated by the statistical model developed using BMARS algorithm. The surrogate model, which is fitted and validated using training dataset generated by the physical model, can approximate solutions rapidly. An efficient Sobol' method is employed to calculate global sensitivities of head outputs to input parameters, which are used to analyze their importance for the model outputs spatiotemporally. Only sensitive parameters are included in the calibration process to further improve the computational efficiency. Normalized root mean square error (NRMSE) between measured and simulated heads at observation wells is used as the objective function to be minimized during optimization. The reasonable history match between the simulated and observed heads demonstrated feasibility of this high-efficient calibration framework.

  13. A Consistent AVHRR Visible Calibration Record Based on Multiple Methods Applicable for the NOAA Degrading Orbits. Part 2 ; Validation

    NASA Technical Reports Server (NTRS)

    Doelling, David R.; Bhatt, Rajendra; Scarino, Benjamin R.; Gopalan, Arun; Haney, Conor O.; Minnis, Patrick; Bedka, Kristopher M.

    2016-01-01

    Consistent cross-sensor Advanced Very High Resolution Radiometer (AVHRR) calibration coefficients are determined using desert, polar ice, and deep convective cloud (DCC) invariant Earth targets. The greatest AVHRR calibration challenge is the slow orbit degradation of the host satellite, which precesses toward a terminator orbit. This issue is solved by characterizing the invariant targets with NOAA-16 AVHRR observed radiances that have been referenced to the Aqua Moderate Resolution Imaging Spectrometer (MODIS) calibration using simultaneous nadir overpass (SNO) observations. Another benefit of the NOAA-16 invariant target-modeled reflectance method is that, because of the similarities among the AVHRR spectral response functions, a smaller spectral band adjustment factor is required than when establishing calibrations relative to a non-AVHRR reference instrument. The sensor- and band-specific calibration uncertainties, with respect to the calibration reference, are, on average, 2 percent and 3 percent for channels 1 and 2, respectively. The uncertainties are smaller for sensors that are in afternoon orbits, have longer records, and spend less time in terminator conditions. The multiple invariant targets referenced to Aqua MODIS (MITRAM) AVHRR calibration coefficients are evaluated for individual target consistency, compared against Aqua MODIS/AVHRR SNOs, and selected published calibration gains. The MITRAM and SNO relative calibration biases mostly agree to within 1 percent for channels 1 and 2, respectively. The individual invariant target and MITRAM sensor relative calibration biases are mostly consistent to within 1 percent and 2 percent for channels 1 and 2, respectively. The differences between the MITRAM and other published calibrations are mostly attributed to the reference instrument calibration differences.

  14. Method for controlling a laser additive process using intrinsic illumination

    NASA Astrophysics Data System (ADS)

    Tait, Robert; Cai, Guoshuang; Azer, Magdi; Chen, Xiaobin; Liu, Yong; Harding, Kevin

    2015-05-01

    One form of additive manufacturing is to use a laser to generate a melt pool from powdered metal that is sprayed from a nozzle. The laser net-shape machining system builds the part a layer at a time by following a predetermined path. However, because the path may need to take many turns, maintaining a constant melt pool may not be easy. A straight section may require one speed and power while a sharp bend would over melt the metal at the same settings. This paper describes a process monitoring method that uses the intrinsic IR radiation from the melt pool along with a process model configured to establish target values for the parameters associated with the manufacture or repair. This model is based upon known properties of the metal being used as well as the properties of the laser beam. An adaptive control technique is then employed to control process parameters of the machining system based upon the real-time weld pool measurement. Since the system uses the heat radiant from the melt pool, other previously deposited metal does not confuse the system as only the melted material is seen by the camera.

  15. Performance of three reflectance calibration methods for airborne hyperspectral spectrometer data.

    PubMed

    Miura, Tomoaki; Huete, Alfredo R

    2009-01-01

    In this study, the performances and accuracies of three methods for converting airborne hyperspectral spectrometer data to reflectance factors were characterized and compared. The "reflectance mode (RM)" method, which calibrates a spectrometer against a white reference panel prior to mounting on an aircraft, resulted in spectral reflectance retrievals that were biased and distorted. The magnitudes of these bias errors and distortions varied significantly, depending on time of day and length of the flight campaign. The "linear-interpolation (LI)" method, which converts airborne spectrometer data by taking a ratio of linearly-interpolated reference values from the preflight and post-flight reference panel readings, resulted in precise, but inaccurate reflectance retrievals. These reflectance spectra were not distorted, but were subject to bias errors of varying magnitudes dependent on the flight duration length. The "continuous panel (CP)" method uses a multi-band radiometer to obtain continuous measurements over a reference panel throughout the flight campaign, in order to adjust the magnitudes of the linear-interpolated reference values from the preflight and post-flight reference panel readings. Airborne hyperspectral reflectance retrievals obtained using this method were found to be the most accurate and reliable reflectance calibration method. The performances of the CP method in retrieving accurate reflectance factors were consistent throughout time of day and for various flight durations. Based on the dataset analyzed in this study, the uncertainty of the CP method has been estimated to be 0.0025 ± 0.0005 reflectance units for the wavelength regions not affected by atmospheric absorptions. The RM method can produce reasonable results only for a very short-term flight (e.g., < 15 minutes) conducted around a local solar noon. The flight duration should be kept shorter than 30 minutes for the LI method to produce results with reasonable accuracies. An important

  16. A Method to Solve Interior and Exterior Camera Calibration Parameters for Image Resection

    NASA Technical Reports Server (NTRS)

    Samtaney, Ravi

    1999-01-01

    An iterative method is presented to solve the internal and external camera calibration parameters, given model target points and their images from one or more camera locations. The direct linear transform formulation was used to obtain a guess for the iterative method, and herein lies one of the strengths of the present method. In all test cases, the method converged to the correct solution. In general, an overdetermined system of nonlinear equations is solved in the least-squares sense. The iterative method presented is based on Newton-Raphson for solving systems of nonlinear algebraic equations. The Jacobian is analytically derived and the pseudo-inverse of the Jacobian is obtained by singular value decomposition.

  17. An Inexpensive Method for Kinematic Calibration of a Parallel Robot by Using One Hand-Held Camera as Main Sensor

    PubMed Central

    Traslosheros, Alberto; Sebastián, José María; Torrijos, Jesús; Carelli, Ricardo; Castillo, Eduardo

    2013-01-01

    This paper presents a novel method for the calibration of a parallel robot, which allows a more accurate configuration instead of a configuration based on nominal parameters. It is used, as the main sensor with one camera installed in the robot hand that determines the relative position of the robot with respect to a spherical object fixed in the working area of the robot. The positions of the end effector are related to the incremental positions of resolvers of the robot motors. A kinematic model of the robot is used to find a new group of parameters, which minimizes errors in the kinematic equations. Additionally, properties of the spherical object and intrinsic camera parameters are utilized to model the projection of the object in the image and thereby improve spatial measurements. Finally, several working tests, static and tracking tests are executed in order to verify how the robotic system behaviour improves by using calibrated parameters against nominal parameters. In order to emphasize that, this proposed new method uses neither external nor expensive sensor. That is why new robots are useful in teaching and research activities. PMID:23921827

  18. Performance of robust regression methods in real-time polymerase chain reaction calibration.

    PubMed

    Orenti, Annalisa; Marubini, Ettore

    2014-12-09

    The ordinary least squares (OLS) method is routinely used to estimate the unknown concentration of nucleic acids in a given solution by means of calibration. However, when outliers are present it could appear sensible to resort to robust regression methods. We analyzed data from an External Quality Control program concerning quantitative real-time PCR and we found that 24 laboratories out of 40 presented outliers, which occurred most frequently at the lowest concentrations. In this article we investigated and compared the performance of the OLS method, the least absolute deviation (LAD) method, and the biweight MM-estimator in real-time PCR calibration via a Monte Carlo simulation. Outliers were introduced by replacement contamination. When contamination was absent the coverages of OLS and MM-estimator intervals were acceptable and their widths small, whereas LAD intervals had acceptable coverages at the expense of higher widths. In the presence of contamination we observed a trade-off between width and coverage: the OLS performance got worse, the MM-estimator intervals widths remained short (but this was associated with a reduction in coverages), while LAD intervals widths were constantly larger with acceptable coverages at the nominal level.

  19. Cloned plasmid DNA fragments as calibrators for controlling GMOs: different real-time duplex quantitative PCR methods.

    PubMed

    Taverniers, Isabel; Van Bockstaele, Erik; De Loose, Marc

    2004-03-01

    Analytical real-time PCR technology is a powerful tool for implementation of the GMO labeling regulations enforced in the EU. The quality of analytical measurement data obtained by quantitative real-time PCR depends on the correct use of calibrator and reference materials (RMs). For GMO methods of analysis, the choice of appropriate RMs is currently under debate. So far, genomic DNA solutions from certified reference materials (CRMs) are most often used as calibrators for GMO quantification by means of real-time PCR. However, due to some intrinsic features of these CRMs, errors may be expected in the estimations of DNA sequence quantities. In this paper, two new real-time PCR methods are presented for Roundup Ready soybean, in which two types of plasmid DNA fragments are used as calibrators. Single-target plasmids (STPs) diluted in a background of genomic DNA were used in the first method. Multiple-target plasmids (MTPs) containing both sequences in one molecule were used as calibrators for the second method. Both methods simultaneously detect a promoter 35S sequence as GMO-specific target and a lectin gene sequence as endogenous reference target in a duplex PCR. For the estimation of relative GMO percentages both "delta C(T)" and "standard curve" approaches are tested. Delta C(T) methods are based on direct comparison of measured C(T) values of both the GMO-specific target and the endogenous target. Standard curve methods measure absolute amounts of target copies or haploid genome equivalents. A duplex delta C(T) method with STP calibrators performed at least as well as a similar method with genomic DNA calibrators from commercial CRMs. Besides this, high quality results were obtained with a standard curve method using MTP calibrators. This paper demonstrates that plasmid DNA molecules containing either one or multiple target sequences form perfect alternative calibrators for GMO quantification and are especially suitable for duplex PCR reactions.

  20. NMR Stark Spectroscopy: New Methods to Calibrate NMR Sensitivity to Electric Fields

    NASA Astrophysics Data System (ADS)

    Tarasek, Matthew R.

    The influence of electrostatics on NMR parameters is well accepted. Thus, NMR is a promising route to probe electrical features within molecules and materials. However, applications of NMR Stark effects (E-field induced changes in spin energy levels) have been elusive. I have developed new approaches to resolve NMR Stark effects from an applied E field. This calibrates nuclear probes whose spectral response might later be used to evaluate internal E fields that are critical to function, such as those due to local charge distributions or sample structure. I will present two novel experimental approaches for direct calibration of NMR quadrupolar Stark effects (QSEs). In the first, steady-state (few-second) excitation by an E field at twice the NMR frequency (2ω 0) is used to saturate spin magnetization. The extent of saturation vs. E-field amplitude calibrates the QSE response rate, while measurements vs sample orientation determine tensorial character. The second method instead synchronizes short (few µs) pulses of the 2ω0 E field with a multiple-pulse NMR sequence. This, “POWER” (Perturbations Observed With Enhanced Resolution) approach enables more accurate measure of small QSEs (i.e. few Hz spectral changes). A 2nd key advantage is the ability to define tensorial response without reorienting the sample, but instead varying the phase of the 2ω0 field. I will describe these experiments and my home-built NMR “Stark probe”, employed on a conventional wide-bore solid-state NMR system. Results with GaAs demonstrate each method, while extensions to a wider array of molecular and material systems may now be possible using these methods.

  1. Investigation of methods for calibration of classifier scores to probability of disease

    NASA Astrophysics Data System (ADS)

    Chen, Weijie; Sahiner, Berkman; Samuelson, Frank; Pezeshk, Aria; Petrick, Nicholas

    2015-03-01

    Classifier scores in many diagnostic devices, such as computer-aided diagnosis systems, are usually on an arbitrary scale, the meaning of which is unclear. Calibration of classifier scores to a meaningful scale such as the probability of disease is potentially useful when such scores are used by a physician or another algorithm. In this work, we investigated the properties of two methods for calibrating classifier scores to probability of disease. The first is a semiparametric method in which the likelihood ratio for each score is estimated based on a semiparametric proper receiver operating characteristic model, and then an estimate of the probability of disease is obtained using the Bayes theorem assuming a known prevalence of disease. The second method is nonparametric in which isotonic regression via the pool-adjacent-violators algorithm is used. We employed the mean square error (MSE) and the Brier score to evaluate the two methods. We evaluate the methods under two paradigms: (a) the dataset used to construct the score-to-probability mapping function is used to calculate the performance metric (MSE or Brier score) (resubstitution); (b) an independent test dataset is used to calculate the performance metric (independent). Under our simulation conditions, the semiparametric method is found to be superior to the nonparametric method at small to medium sample sizes and the two methods appear to converge at large sample sizes. Our simulation results also indicate that the resubstitution bias may depend on the performance metric and, for the semiparametric method, the resubstitution bias is small when a reasonable number of cases (> 100 cases per class) are available.

  2. Bivariate versus multivariate smart spectrophotometric calibration methods for the simultaneous determination of a quaternary mixture of mosapride, pantoprazole and their degradation products.

    PubMed

    Hegazy, M A; Yehia, A M; Moustafa, A A

    2013-05-01

    The ability of bivariate and multivariate spectrophotometric methods was demonstrated in the resolution of a quaternary mixture of mosapride, pantoprazole and their degradation products. The bivariate calibrations include bivariate spectrophotometric method (BSM) and H-point standard addition method (HPSAM), which were able to determine the two drugs, simultaneously, but not in the presence of their degradation products, the results showed that simultaneous determinations could be performed in the concentration ranges of 5.0-50.0 microg/ml for mosapride and 10.0-40.0 microg/ml for pantoprazole by bivariate spectrophotometric method and in the concentration ranges of 5.0-45.0 microg/ml for both drugs by H-point standard addition method. Moreover, the applied multivariate calibration methods were able for the determination of mosapride, pantoprazole and their degradation products using concentration residuals augmented classical least squares (CRACLS) and partial least squares (PLS). The proposed multivariate methods were applied to 17 synthetic samples in the concentration ranges of 3.0-12.0 microg/ml mosapride, 8.0-32.0 microg/ml pantoprazole, 1.5-6.0 microg/ml mosapride degradation products and 2.0-8.0 microg/ml pantoprazole degradation products. The proposed bivariate and multivariate calibration methods were successfully applied to the determination of mosapride and pantoprazole in their pharmaceutical preparations.

  3. Linear Calibration of Radiographic Mineral Density Using Video-Digitizing Methods

    NASA Technical Reports Server (NTRS)

    Martin, R. Bruce; Papamichos, Thomas; Dannucci, Greg A.

    1990-01-01

    Radiographic images can provide quantitative as well as qualitative information if they are subjected to densitometric analysis. Using modern video-digitizing techniques, such densitometry can be readily accomplished using relatively inexpensive computer systems. However, such analyses are made more difficult by the fact that the density values read from the radiograph have a complex, nonlinear relationship to bone mineral content. This article derives the relationship between these variables from the nature of the intermediate physical processes, and presents a simple mathematical method for obtaining a linear calibration function using a step wedge or other standard.

  4. Linear Calibration of Radiographic Mineral Density Using Video-Digitizing Methods

    NASA Technical Reports Server (NTRS)

    Martin, R. Bruce; Papamichos, Thomas; Dannucci, Greg A.

    1990-01-01

    Radiographic images can provide quantitative as well as qualitative information if they are subjected to densitometric analysis. Using modem video-digitizing techniques, such densitometry can be readily accomplished using relatively inexpensive computer systems. However, such analyses are made more difficult by the fact that the density values read from the radiograph have a complex, nonlinear relationship to bone mineral content. This article derives the relationship between these variables from the nature of the intermediate physical processes, and presents a simple mathematical method for obtaining a linear calibration function using a step wedge or other standard.

  5. Towards a shock tube method for the dynamic calibration of pressure sensors

    PubMed Central

    Downes, Stephen; Knott, Andy; Robinson, Ian

    2014-01-01

    In theory, shock tubes provide a pressure change with a very fast rise time and calculable amplitude. This pressure step could provide the basis for the calibration of pressure transducers used in highly dynamic applications. However, conventional metal shock tubes can be expensive, unwieldy and difficult to modify. We describe the development of a 1.4 MPa (maximum pressure) shock tube made from unplasticized polyvinyl chloride pressure tubing which provides a low-cost, light and easily modifiable basis for establishing a method for determining the dynamic characteristics of pressure sensors. PMID:25071242

  6. The New Calibration Method of Accelerometer in GRACE Satellites Based on Precise Solar Radiation Model

    NASA Astrophysics Data System (ADS)

    Wang, H. B.; Xiong, Y. Q.; Zhao, C. Y.

    2016-09-01

    In this paper, we adopt the precise solar radiation model to compute the real perturbation force caused by solar radiation on the GRACE satellites, and estimate the scale factors of accelerometer's y-axis and z-axis, which are set to be constant values in the following calibration process. Then, we estimate the rest of four parameters by the dynamic orbit determination (OD), such as the scale factor of x-axis, and the biases of three axes. Through these steps, we get the daily calibration parameters from 2002 to 2014. The average value and standard deviation of scale factors of x-, y-, and z-axis are 0.9435± 0.0187, 0.9393± 0.0444, 1.0371± 0.0391 for GRACE-A, and 0.9313± 0.0170, 0.9488± 0.0452, 1.0274± 0.0446 for GRACE-B, respectively. Different from our early work, the new method constrains the scale factors of y-axis and z-axis with the precise solar radiation model, which could decrease the measurement error's effect on the weak-signal axes (y, z) as well as reduce the correlation between scale factor and bias, and eventually improve the stability of calibration parameters. Taking the results of y- and z-axis of GRACE-A as example, the standard deviation of scale factors with the new method is about 0.0391-0.0444, while the early results by the unconstrained dynamic orbit determination are about 0.21-0.31. It is shown that the standard deviations of scale factor with this paper's method have been decreased by more than 78%, and those of bias have been decreased by more than 85%. Therefore, the calibration parameters estimated with the new method are more stable than our early results, and will have a special value for the study on the rotation speed and wind field of the earth's thermosphere.

  7. Improved phase-ellipse method for in-situ geophone calibration.

    USGS Publications Warehouse

    Liu, Huaibao P.; Peselnick, L.

    1986-01-01

    For amplitude and phase response calibration of moving-coil electromagnetic geophones 2 parameters are needed, namely the geophone natural frequency, fo, and the geophone upper resonance frequency fu. The phase-ellipse method is commonly used for the in situ determination of these parameters. For a given signal-to-noise ratio, the precision of the measurement of fo and fu depends on the phase sensitivity, f(delta PHI/delta PHIf). For some commercial geophones (f(delta PHI/delta PHI) at fu can be an order of magnitude less than the sensitivity at fo. Presents an improved phase-ellipse method with increased precision. Compared to measurements made with the existing phase-ellipse methods, the method shows a 6- and 3-fold improvement in the precision, respectively, on measurements of fo and fu on a commercial geophone.-from Authors

  8. A New Method to Calibrate Attachment Angles of Data Loggers in Swimming Sharks

    NASA Astrophysics Data System (ADS)

    Kawatsu, Shizuka; Sato, Katsufumi; Watanabe, Yuuki; Hyodo, Susumu; Breves, Jason P.; Fox, Bradley K.; Grau, E. Gordon; Miyazaki, Nobuyuki

    2009-12-01

    Recently, animal-borne accelerometers have been used to record the pitch angle of aquatic animals during swimming. When evaluating pitch angle, it is necessary to consider a discrepancy between the angle of an accelerometer and the long axis of an animal. In this study, we attached accelerometers to 17 free-ranging scalloped hammerhead shark ( Sphyrna lewini) pups from Kaneohe Bay, Hawaii. Although there are methods to calibrate attachment angles of accelerometers, we confirmed that previous methods were not applicable for hammerhead pups. According to raw data, some sharks ascended with a negative angle, which differs from tank observations of captive sharks. In turn, we developed a new method to account for this discrepancy in swimming sharks by estimating the attachment angle from the relationship between vertical speed (m/s) and pitch angle obtained by each accelerometer. The new method can be utilized for field observation of a wide range of species.

  9. Locally-calibrated light transmission visualization methods to quantify nonaqueous phase liquid mass in porous media

    NASA Astrophysics Data System (ADS)

    Wang, Huaguo; Chen, Xiaosong; Jawitz, James W.

    2008-11-01

    Five locally-calibrated light transmission visualization (LTV) methods were tested to quantify nonaqueous phase liquid (NAPL) mass and mass reduction in porous media. Tetrachloroethylene (PCE) was released into a two-dimensional laboratory flow chamber packed with water-saturated sand which was then flushed with a surfactant solution (2% Tween 80) until all of the PCE had been dissolved. In all the LTV methods employed here, the water phase was dyed, rather than the more common approach of dyeing the NAPL phase, such that the light adsorption characteristics of NAPL did not change as dissolution progressed. Also, none of the methods used here required the use of external calibration chambers. The five visualization approaches evaluated included three methods developed from previously published models, a binary method, and a novel multiple wavelength method that has the advantage of not requiring any assumptions about the intra-pore interface structure between the various phases (sand/water/NAPL). The new multiple wavelength method is also expected to be applicable to any translucent porous media containing two immiscible fluids (e.g., water-air, water-NAPL). Results from the sand-water-PCE system evaluated here showed that the model that assumes wetting media of uniform pore size (Model C of Niemet and Selker, 2001) and the multiple wavelength model with no interface structure assumptions were able to accurately quantify PCE mass reduction during surfactant flushing. The average mass recoveries from these two imaging methods were greater than 95% for domain-average NAPL saturations of approximately 2.6 × 10 - 2 , and were approximately 90% during seven cycles of surfactant flushing that sequentially reduced the average NAPL saturation to 7.5 × 10 - 4 .

  10. Calibration of pulsed electroacoustic method considering electrode-dielectric interface status and porosity

    NASA Astrophysics Data System (ADS)

    Huang, Meng; Zhou, Yuanxiang; Chen, Weijiang; Lu, Licheng; Jin, Fubao; Huang, Jianwen

    2014-10-01

    Pulsed electroacoustic (PEA) method is widely used, but measurement conditions not always agree with the underlying PEA assumptions. This necessitates space charge recovery; however, existing research only addresses the attenuation and dispersion in lossy media. The effects of electrode-dielectric interface contact status and porosity on the accuracy of space charge distribution are discussed in the present article. It is shown that the presence of nonlinear interface and porosity can introduce severe error. However, because the properties of acoustic propagation of waves (which are generated from charge on the electrode and in the bulk) are different, the conventional recovery algorithm is no longer suitable for calibrating the charge density. To obtain accurate space charge profiles, it is necessary to eliminate these effects. A method has been proposed which is based on the original measurement process. The validity of the proposed method was tested by reasonable post-recovery electric field distributions.

  11. Laser ablation methods for analysis of urinary calculi: Comparison study based on calibration pellets

    NASA Astrophysics Data System (ADS)

    Štěpánková, K.; Novotný, K.; Vašinová Galiová, M.; Kanický, V.; Kaiser, J.; Hahn, D. W.

    2013-03-01

    Methods based on laser ablation, such as Laser-Induced Breakdown Spectroscopy (LIBS) and Laser-Ablation Inductively Coupled Plasma Mass/Optical Emission Spectrometry (LA-ICP-MS/OES) are particularly suitable for urinary calculi bulk and micro analysis. Investigation of spatial distribution of matrix and trace elements can help to explain their emergence and growth. However, quantification is still very problematic and these methods are often used only for qualitative elemental mapping. There are no commercially available standards, which would correspond to the urinary calculi matrix. Internal standardization is also difficult, mainly due to different crystalline phases in one kidney stone. The aim of this study is to demonstrate the calibration capabilities and examine the limitations of laser ablation based techniques. Calibration pellets were prepared from powdered human urinary calculi with phosphate, oxalate and urate matrix. For this comparative study, the most frequently used laser-ablation based analytical techniques were chosen, such as LIBS and LA-ICP-MS. Moreover, some alternative techniques such as simultaneous LIBS-LA-ICP-OES and laser ablation LA-LIBS were also utilized.

  12. Self-calibration method of the bias of a space electrostatic accelerometer.

    PubMed

    Qu, Shao-Bo; Xia, Xiao-Mei; Bai, Yan-Zheng; Wu, Shu-Chao; Zhou, Ze-Bing

    2016-11-01

    The high precision space electrostatic accelerometer is an instrument to measure the non-gravitational forces acting on a spacecraft. It is one of the key payloads for satellite gravity measurements and space fundamental physics experiments. The measurement error of the accelerometer directly affects the precision of gravity field recovery for the earth. This paper analyzes the sources of the bias according to the operating principle and structural constitution of the space electrostatic accelerometer. Models of bias due to the asymmetry of the displacement sensing system, including the mechanical sensor head and the capacitance sensing circuit, and the asymmetry of the feedback control actuator circuit are described separately. According to the two models, a method of bias self-calibration by using only the accelerometer data is proposed, based on the feedback voltage data of the accelerometer before and after modulating the DC biasing voltage (Vb) applied on its test mass. Two types of accelerometer biases are evaluated separately using in-orbit measurement data of a space electrostatic accelerometer. Based on the preliminary analysis, the bias of the accelerometer onboard of an experiment satellite is evaluated to be around 10(-4) m/s(2), about 4 orders of magnitude greater than the noise limit. Finally, considering the two asymmetries, a comprehensive bias model is analyzed. A modified method to directly calibrate the accelerometer comprehensive bias is proposed.

  13. Self-calibration method of the bias of a space electrostatic accelerometer

    NASA Astrophysics Data System (ADS)

    Qu, Shao-Bo; Xia, Xiao-Mei; Bai, Yan-Zheng; Wu, Shu-Chao; Zhou, Ze-Bing

    2016-11-01

    The high precision space electrostatic accelerometer is an instrument to measure the non-gravitational forces acting on a spacecraft. It is one of the key payloads for satellite gravity measurements and space fundamental physics experiments. The measurement error of the accelerometer directly affects the precision of gravity field recovery for the earth. This paper analyzes the sources of the bias according to the operating principle and structural constitution of the space electrostatic accelerometer. Models of bias due to the asymmetry of the displacement sensing system, including the mechanical sensor head and the capacitance sensing circuit, and the asymmetry of the feedback control actuator circuit are described separately. According to the two models, a method of bias self-calibration by using only the accelerometer data is proposed, based on the feedback voltage data of the accelerometer before and after modulating the DC biasing voltage (Vb) applied on its test mass. Two types of accelerometer biases are evaluated separately using in-orbit measurement data of a space electrostatic accelerometer. Based on the preliminary analysis, the bias of the accelerometer onboard of an experiment satellite is evaluated to be around 10-4 m/s2, about 4 orders of magnitude greater than the noise limit. Finally, considering the two asymmetries, a comprehensive bias model is analyzed. A modified method to directly calibrate the accelerometer comprehensive bias is proposed.

  14. Novel method of calibration with restrictive constraints for stereo-vision system

    NASA Astrophysics Data System (ADS)

    Cui, Jiashan; Huo, Ju; Yang, Ming

    2016-05-01

    Regarding the calibration of a stereo vision measurement system, this paper puts forward a new bundle adjustment algorithm based on the stereo vision camera calibration method. Multiple-view geometric constraints and a bundle adjustment algorithm are used to optimize the inner and outer parameters of the camera accurately. A fixed relative constraint relationship between cameras is introduced. We have improved the normal equation construction process of the traditional bundle adjustment method, so that each iteration process occurs just outside the parameters of two images that are taken by a camera that has been optimized to better integrate two cameras bound together as one camera. The relationship between the fixed relative constraints can effectively increase the number of superfluous observations of the adjustment system and optimize higher accuracy while reducing the dimension of the normal matrix; it means that each iteration will reduce the time required. Simulation and actual experimental results show the superior performance of the proposed approach in terms of robustness and accuracy, and our approach also can be extended to stereo-vision system with more than two cameras.

  15. A new data processing and calibration method for an eye-tracking device pronunciation system

    NASA Astrophysics Data System (ADS)

    Lin, Chern-Sheng; Chang, Kai-Chieh; Jain, Young-Jou

    2002-07-01

    In this paper, a new data processing and calibration method for a pronunciation system of an eye-tracking device is described. The eye-tracking device was created using both head mounted display (HMD) technology and remote operation capabilities. A pattern recognition computer program was used to distinguish the pupil position and calculate its coordinates. This system can be adapted to provide a digital speech function. A new method for processing the image of the eye in the PC-based system was also developed. With one video CCD camera and frame grabber analyzing a series of human pupil images while the subject is gazing at the screen, an auto-calibration algorithm is used to obtain the direction of the eye gaze in real time. The computers provide the speech sound according to the location where the eye gazes exceed 0.5 s. The availability of multipurpose in this eye-tracking system with very simple equipment will be reconfirmed for future advanced research.

  16. Methods for Calibrating Basin-Wide Hydroacoustic Propagation in the Indian Ocean

    SciTech Connect

    Blackman, D; de Groot-Hedlin, C; Orcutt, J A; Harben, P H; Clarke, D B; Ramirez, A L

    2004-10-11

    This collaborative project was designed to test and compare methods for achieving full ocean basin propagation of hydroacoustic signals in the 5-100 Hz frequency band. Plans for a systematic calibration of the International Monitoring System (IMS) for nuclear testing were under consideration in 2000/2001. The results from this project provide information to guide such planning for future ocean basin calibration work. Several acoustic source types were tested during two sea-going experiments and most were successful at generating signals that propagated hundreds to thousands of km to be recorded at the Indian Ocean IMS hydrophone stations. Development and numerical modeling of imploding glass sphere sources was one component of this testing. The intent was to design a relatively simple-to-use source that is not subject to restrictions that can limit use of explosive charges, but whose signal is large enough to propagate 100-1000's km range. Analysis of IMS hydrophone data recording during the experiments was used to illustrate the extent of energy loss during signal propagation and to assess the accuracy with which the small acoustic sources could be located using methods typically employed for nuclear monitoring.

  17. Additives and method for controlling clathrate hydrates in fluid systems

    DOEpatents

    Sloan, Jr., Earle Dendy; Christiansen, Richard Lee; Lederhos, Joseph P.; Long, Jin Ping; Panchalingam, Vaithilingam; Du, Yahe; Sum, Amadeu Kun Wan

    1997-01-01

    Discussed is a process for preventing clathrate hydrate masses from detrimentally impeding the possible flow of a fluid susceptible to clathrate hydrate formation. The process is particularly useful in the natural gas and petroleum production, transportation and processing industry where gas hydrate formation can cause serious problems. Additives preferably contain one or more five member, six member and/or seven member cyclic chemical groupings. Additives include polymers having lactam rings. Additives can also contain polyelectrolytes that are believed to improve conformance of polymer additives through steric hinderance and/or charge repulsion. Also, polymers having an amide on which a C.sub.1 -C.sub.4 group is attached to the nitrogen and/or the carbonyl carbon of the amide may be used alone, or in combination with ring-containing polymers for enhanced effectiveness. Polymers having at least some repeating units representative of polymerizing at least one of an oxazoline, an N-substituted acrylamide and an N-vinyl alkyl amide are preferred.

  18. Additives and method for controlling clathrate hydrates in fluid systems

    DOEpatents

    Sloan, E.D. Jr.; Christiansen, R.L.; Lederhos, J.P.; Long, J.P.; Panchalingam, V.; Du, Y.; Sum, A.K.W.

    1997-06-17

    Discussed is a process for preventing clathrate hydrate masses from detrimentally impeding the possible flow of a fluid susceptible to clathrate hydrate formation. The process is particularly useful in the natural gas and petroleum production, transportation and processing industry where gas hydrate formation can cause serious problems. Additives preferably contain one or more five member, six member and/or seven member cyclic chemical groupings. Additives include polymers having lactam rings. Additives can also contain polyelectrolytes that are believed to improve conformance of polymer additives through steric hindrance and/or charge repulsion. Also, polymers having an amide on which a C{sub 1}-C{sub 4} group is attached to the nitrogen and/or the carbonyl carbon of the amide may be used alone, or in combination with ring-containing polymers for enhanced effectiveness. Polymers having at least some repeating units representative of polymerizing at least one of an oxazoline, an N-substituted acrylamide and an N-vinyl alkyl amide are preferred.

  19. Online Calibration Methods for the DINA Model with Independent Attributes in CD-CAT

    ERIC Educational Resources Information Center

    Chen, Ping; Xin, Tao; Wang, Chun; Chang, Hua-Hua

    2012-01-01

    Item replenishing is essential for item bank maintenance in cognitive diagnostic computerized adaptive testing (CD-CAT). In regular CAT, online calibration is commonly used to calibrate the new items continuously. However, until now no reference has publicly become available about online calibration for CD-CAT. Thus, this study investigates the…

  20. A novel functional calibration method for real-time elbow joint angles estimation with magnetic-inertial sensors.

    PubMed

    Ligorio, G; Zanotto, D; Sabatini, A M; Agrawal, S K

    2017-01-30

    Magnetic-inertial measurement units (MIMUs) are often used to measure the joint angles between two body segments. To obtain anatomically meaningful joint angles, each MIMU must be computationally aligned (i.e., calibrated) with the anatomical rotation axes. In this paper, a novel four-step functional calibration method is presented for the elbow joint, which relies on a two-degrees-of-freedom elbow model. In each step, subjects are asked to perform a simple task involving either one-dimensional motions around some anatomical axes or a static posture. The proposed method was implemented on a fully portable wearable system, which, after calibration, was capable of estimating the elbow joint angles in real time. Fifteen subjects participated in a multi-session experiment that was designed to assess accuracy, repeatability and robustness of the proposed method. When compared against an optical motion capture system (OMCS), the proposed wearable system showed an accuracy of about 4° along each degree of freedom. The proposed calibration method was tested against different MIMU mountings, multiple repetitions and non-strict observance of the calibration protocol and proved to be robust against these factors. Compared to previous works, the proposed method does not require the wearer to maintain specific arm postures while performing the calibration motions, and therefore it is more robust and better suited for real-world applications.

  1. Ionospheric calibration of low frequency radio interferometric observations using the peeling scheme. I. Method description and first results

    NASA Astrophysics Data System (ADS)

    Intema, H. T.; van der Tol, S.; Cotton, W. D.; Cohen, A. S.; van Bemmel, I. M.; Röttgering, H. J. A.

    2009-07-01

    Calibration of radio interferometric observations becomes increasingly difficult towards lower frequencies. Below ˜ 300 MHz, spatially variant refractions and propagation delays of radio waves traveling through the ionosphere cause phase rotations that can vary significantly with time, viewing direction and antenna location. In this article we present a description and first results of SPAM (Source Peeling and Atmospheric Modeling), a new calibration method that attempts to iteratively solve and correct for ionospheric phase errors. To model the ionosphere, we construct a time-variant, 2-dimensional phase screen at fixed height above the Earth's surface. Spatial variations are described by a truncated set of discrete Karhunen-Loève base functions, optimized for an assumed power-law spectral density of free electrons density fluctuations, and a given configuration of calibrator sources and antenna locations. The model is constrained using antenna-based gain phases from individual self-calibrations on the available bright sources in the field-of-view. Application of SPAM on three test cases, a simulated visibility data set and two selected 74 MHz VLA data sets, yields significant improvements in image background noise (5-75 percent reduction) and source peak fluxes (up to 25 percent increase) as compared to the existing self-calibration and field-based calibration methods, which indicates a significant improvement in ionospheric phase calibration accuracy.

  2. Development and validation of multivariate calibration methods for simultaneous estimation of Paracetamol, Enalapril maleate and hydrochlorothiazide in pharmaceutical dosage form

    NASA Astrophysics Data System (ADS)

    Singh, Veena D.; Daharwal, Sanjay J.

    2017-01-01

    Three multivariate calibration spectrophotometric methods were developed for simultaneous estimation of Paracetamol (PARA), Enalapril maleate (ENM) and Hydrochlorothiazide (HCTZ) in tablet dosage form; namely multi-linear regression calibration (MLRC), trilinear regression calibration method (TLRC) and classical least square (CLS) method. The selectivity of the proposed methods were studied by analyzing the laboratory prepared ternary mixture and successfully applied in their combined dosage form. The proposed methods were validated as per ICH guidelines and good accuracy; precision and specificity were confirmed within the concentration range of 5-35 μg mL- 1, 5-40 μg mL- 1 and 5-40 μg mL- 1of PARA, HCTZ and ENM, respectively. The results were statistically compared with reported HPLC method. Thus, the proposed methods can be effectively useful for the routine quality control analysis of these drugs in commercial tablet dosage form.

  3. Comparison of function approximation, heuristic, and derivative-based methods for automatic calibration of computationally expensive groundwater bioremediation models

    NASA Astrophysics Data System (ADS)

    Mugunthan, Pradeep; Shoemaker, Christine A.; Regis, Rommel G.

    2005-11-01

    The performance of function approximation (FA) methods is compared to heuristic and derivative-based nonlinear optimization methods for automatic calibration of biokinetic parameters of a groundwater bioremediation model of chlorinated ethenes on a hypothetical and a real field case. For the hypothetical case, on the basis of 10 trials on two different objective functions, the FA methods had the lowest mean and smaller deviation of the objective function among all algorithms for a combined Nash-Sutcliffe objective and among all but the derivative-based algorithm for a total squared error objective. The best algorithms in the hypothetical case were applied to calibrate eight parameters to data obtained from a site in California. In three trials the FA methods outperformed heuristic and derivative-based methods for both objective functions. This study indicates that function approximation methods could be a more efficient alternative to heuristic and derivative-based methods for automatic calibration of computationally expensive bioremediation models.

  4. A Measuring System for Well Logging Attitude and a Method of Sensor Calibration

    PubMed Central

    Ren, Yong; Wang, Yangdong; Wang, Mijian; Wu, Sheng; Wei, Biao

    2014-01-01

    This paper proposes an approach for measuring the azimuth angle and tilt angle of underground drilling tools with a MEMS three-axis accelerometer and a three-axis fluxgate sensor. A mathematical model of well logging attitude angle is deduced based on combining space coordinate transformations and algebraic equations. In addition, a system implementation plan of the inclinometer is given in this paper, which features low cost, small volume and integration. Aiming at the sensor and assembly errors, this paper analyses the sources of errors, and establishes two mathematical models of errors and calculates related parameters to achieve sensor calibration. The results show that this scheme can obtain a stable and high precision azimuth angle and tilt angle of drilling tools, with the deviation of the former less than ±1.4° and the deviation of the latter less than ±0.1°. PMID:24859028

  5. A measuring system for well logging attitude and a method of sensor calibration.

    PubMed

    Ren, Yong; Wang, Yangdong; Wang, Mijian; Wu, Sheng; Wei, Biao

    2014-05-23

    This paper proposes an approach for measuring the azimuth angle and tilt angle of underground drilling tools with a MEMS three-axis accelerometer and a three-axis fluxgate sensor. A mathematical model of well logging attitude angle is deduced based on combining space coordinate transformations and algebraic equations. In addition, a system implementation plan of the inclinometer is given in this paper, which features low cost, small volume and integration. Aiming at the sensor and assembly errors, this paper analyses the sources of errors, and establishes two mathematical models of errors and calculates related parameters to achieve sensor calibration. The results show that this scheme can obtain a stable and high precision azimuth angle and tilt angle of drilling tools, with the deviation of the former less than ±1.4° and the deviation of the latter less than ±0.1°.

  6. Gasoline and Diesel Fuel Test Methods Additional Resources

    EPA Pesticide Factsheets

    Supporting documents on the Direct Final Rule that allows refiners and laboratories to use more current and improved fuel testing procedures for twelve American Society for Testing and Materials analytical test methods.

  7. Simultaneous analysis of riboflavin and aromatic amino acids in beer using fluorescence and multivariate calibration methods.

    PubMed

    Sikorska, Ewa; Gliszczyńska-Swigło, Anna; Insińska-Rak, Małgorzata; Khmelinskii, Igor; De Keukeleire, Denis; Sikorski, Marek

    2008-04-21

    The study demonstrates an application of the front-face fluorescence spectroscopy combined with multivariate regression methods to the analysis of fluorescent beer components. Partial least-squares regressions (PLS1, PLS2, and N-way PLS) were utilized to develop calibration models between synchronous fluorescence spectra and excitation-emission matrices of beers, on one hand, and analytical concentrations of riboflavin and aromatic amino acids, on the other hand. The best results were obtained in the analysis of excitation-emission matrices using the N-way PLS2 method. The respective correlation coefficients, and the values of the root mean-square error of cross-validation (RMSECV), expressed as percentages of the respective mean analytic concentrations, were: 0.963 and 14% for riboflavin, 0.974 and 4% for tryptophan, 0.980 and 4% for tyrosine, and 0.982 and 19% for phenylalanine.

  8. Method for in-situ restoration of platinum resistance thermometer calibration

    DOEpatents

    Carroll, R.M.

    1987-10-23

    A method is provided for in-situ restoration of platinum resistance thermometers (PRT's) that have undergone surface oxide contamination and/or stain-related damage causing decalibration. The method, which may be automated using a programmed computer control arrangement, consists of applying a dc heating current to the resistive sensing element of the PRT of sufficient magnitude to heat the element to an annealing temperature and maintaining the temperature for a specified period to restore the element to a stress-free calibration condition. The process anneals the sensing element of the PRT without subjecting the entire PRT assembly to the annealing temperature and may be used in the periodic maintenance of installed PRT's. 1 fig.

  9. Method for in-situ restoration of plantinum resistance thermometer calibration

    DOEpatents

    Carroll, Radford M.

    1989-01-01

    A method is provided for in-situ restoration of platinum resistance thermometers (PRT's) that have undergone surface oxide contamination and/or strain-related damage causing decalibration. The method, which may be automated using a programmed computer control arrangement, consists of applying a dc heating current to the resistive sensing element of the PRT of sufficient magnitude to heat the element to an annealing temperature and maintaining the temperature for a specified period to restore the element to a stress-free calibration condition. The process anneals the sensing element of the PRT without subjecting the entire PRT assembly to the annealing temperature and may be used in the periodic maintenance of installed PRT's.

  10. A calibration method for measurement of small alkyl organic peroxy radicals by chemical amplifier.

    PubMed

    Qi, Bin; Takami, Akinori; Hatakeyama, Shiro

    2006-08-01

    A new method is proposed to determine the calibration factor (CF) of methyl and ethyl peroxy radicals in a chemical amplifier. The radical source comes from the reactions of excess methane and ethane, respectively, with known concentrations of OH radicals generated by the photolysis of water vapor at 184.9 nm in air in a flow tube. This yields a mixed radical source with equal amounts of HO2 and RO2 (R = CH3, C2H5). The CF for RO2 can be derived from the CF for HO2 and an average CF for the mixed radicals. The reliability of the method was evaluated by comparing the CF ratios of RO2 to HO2 obtained from both the experiments and theoretical calculations.

  11. A Novel Error Model of Optical Systems and an On-Orbit Calibration Method for Star Sensors

    PubMed Central

    Wang, Shuang; Geng, Yunhai; Jin, Rongyu

    2015-01-01

    In order to improve the on-orbit measurement accuracy of star sensors, the effects of image-plane rotary error, image-plane tilt error and distortions of optical systems resulting from the on-orbit thermal environment were studied in this paper. Since these issues will affect the precision of star image point positions, in this paper, a novel measurement error model based on the traditional error model is explored. Due to the orthonormal characteristics of image-plane rotary-tilt errors and the strong nonlinearity among these error parameters, it is difficult to calibrate all the parameters simultaneously. To solve this difficulty, for the new error model, a modified two-step calibration method based on the Extended Kalman Filter (EKF) and Least Square Methods (LSM) is presented. The former one is used to calibrate the main point drift, focal length error and distortions of optical systems while the latter estimates the image-plane rotary-tilt errors. With this calibration method, the precision of star image point position influenced by the above errors is greatly improved from 15.42% to 1.389%. Finally, the simulation results demonstrate that the presented measurement error model for star sensors has higher precision. Moreover, the proposed two-step method can effectively calibrate model error parameters, and the calibration precision of on-orbit star sensors is also improved obviously. PMID:26703599

  12. A Novel Error Model of Optical Systems and an On-Orbit Calibration Method for Star Sensors.

    PubMed

    Wang, Shuang; Geng, Yunhai; Jin, Rongyu

    2015-12-12

    In order to improve the on-orbit measurement accuracy of star sensors, the effects of image-plane rotary error, image-plane tilt error and distortions of optical systems resulting from the on-orbit thermal environment were studied in this paper. Since these issues will affect the precision of star image point positions, in this paper, a novel measurement error model based on the traditional error model is explored. Due to the orthonormal characteristics of image-plane rotary-tilt errors and the strong nonlinearity among these error parameters, it is difficult to calibrate all the parameters simultaneously. To solve this difficulty, for the new error model, a modified two-step calibration method based on the Extended Kalman Filter (EKF) and Least Square Methods (LSM) is presented. The former one is used to calibrate the main point drift, focal length error and distortions of optical systems while the latter estimates the image-plane rotary-tilt errors. With this calibration method, the precision of star image point position influenced by the above errors is greatly improved from 15.42% to 1.389%. Finally, the simulation results demonstrate that the presented measurement error model for star sensors has higher precision. Moreover, the proposed two-step method can effectively calibrate model error parameters, and the calibration precision of on-orbit star sensors is also improved obviously.

  13. Sensitivity and Calibration of Non-Destructive Evaluation Method That Uses Neural-Net Processing of Characteristic Fringe Patterns

    NASA Technical Reports Server (NTRS)

    Decker, Arthur J.; Weiland, Kenneth E.

    2003-01-01

    This paper answers some performance and calibration questions about a non-destructive-evaluation (NDE) procedure that uses artificial neural networks to detect structural damage or other changes from sub-sampled characteristic patterns. The method shows increasing sensitivity as the number of sub-samples increases from 108 to 6912. The sensitivity of this robust NDE method is not affected by noisy excitations of the first vibration mode. A calibration procedure is proposed and demonstrated where the output of a trained net can be correlated with the outputs of the point sensors used for vibration testing. The calibration procedure is based on controlled changes of fastener torques. A heterodyne interferometer is used as a displacement sensor for a demonstration of the challenges to be handled in using standard point sensors for calibration.

  14. A calibrator based on the use of low-coherent light source straightness interferometer and compensation method.

    PubMed

    Lin, Shyh-Tsong; Yeh, Sheng-Lih; Chiu, Chi-Shang; Huang, Mou-Shan

    2011-10-24

    A calibrator utilizing a low-coherent light source straightness interferometer and a compensation method is introduced for straightness measurements in this paper. Where the interference pattern, which is modulated by an envelope function, generated by the interferometer undergoes a shifting as the Wolaston prism of the interferometer experiences a lateral displacement, and the compensation method senses the displacement by driving the prism back to the position to restore the pattern. A setup, which is with a measurement sensitivity of 36.6°/μm, constructed for realizing the calibrator is demonstrated. The experimental results from the uses of the setup reveal that the setup is with a measurement resolution and stability of 0.019 and 0.08 μm, respectively, validate the calibrator, and confirm the calibrator's applicability of straightness measurements and advantage of extensible working distance.

  15. A validated method for the quantitation of 1,1-difluoroethane using a gas in equilibrium method of calibration.

    PubMed

    Avella, Joseph; Lehrer, Michael; Zito, S William

    2008-10-01

    1,1-Difluoroethane (DFE), also known as Freon 152A, is a member of a class of compounds known as halogenated hydrocarbons. A number of these compounds have gained notoriety because of their ability to induce rapid onset of intoxication after inhalation exposure. Abuse of DFE has necessitated development of methods for its detection and quantitation in postmortem and human performance specimens. Furthermore, methodologies applicable to research studies are required as there have been limited toxicokinetic and toxicodynamic reports published on DFE. This paper describes a method for the quantitation of DFE using a gas chromatography-flame-ionization headspace technique that employs solventless standards for calibration. Two calibration curves using 0.5 mL whole blood calibrators which ranged from A: 0.225-1.350 to B: 9.0-180.0 mg/L were developed. These were evaluated for linearity (0.9992 and 0.9995), limit of detection of 0.018 mg/L, limit of quantitation of 0.099 mg/L (recovery 111.9%, CV 9.92%), and upper limit of linearity of 27,000.0 mg/L. Combined curve recovery results of a 98.0 mg/L DFE control that was prepared using an alternate technique was 102.2% with CV of 3.09%. No matrix interference was observed in DFE enriched blood, urine or brain specimens nor did analysis of variance detect any significant differences (alpha = 0.01) in the area under the curve of blood, urine or brain specimens at three identical DFE concentrations. The method is suitable for use in forensic laboratories because validation was performed on instrumentation routinely used in forensic labs and due to the ease with which the calibration range can be adjusted. Perhaps more importantly it is also useful for research oriented studies because the removal of solvent from standard preparation eliminates the possibility for solvent induced changes to the gas/liquid partitioning of DFE or chromatographic interference due to the presence of solvent in specimens.

  16. Quasi-Static Calibration Method of a High-g Accelerometer.

    PubMed

    Wang, Yan; Fan, Jinbiao; Zu, Jing; Xu, Peng

    2017-02-20

    To solve the problem of resonance during quasi-static calibration of high-g accelerometers, we deduce the relationship between the minimum excitation pulse width and the resonant frequency of the calibrated accelerometer according to the second-order mathematical model of the accelerometer, and improve the quasi-static calibration theory. We establish a quasi-static calibration testing system, which uses a gas gun to generate high-g acceleration signals, and apply a laser interferometer to reproduce the impact acceleration. These signals are used to drive the calibrated accelerometer. By comparing the excitation acceleration signal and the output responses of the calibrated accelerometer to the excitation signals, the impact sensitivity of the calibrated accelerometer is obtained. As indicated by the calibration test results, this calibration system produces excitation acceleration signals with a pulse width of less than 1000 μs, and realize the quasi-static calibration of high-g accelerometers with a resonant frequency above 20 kHz when the calibration error was 3%.

  17. Quasi-Static Calibration Method of a High-g Accelerometer

    PubMed Central

    Wang, Yan; Fan, Jinbiao; Zu, Jing; Xu, Peng

    2017-01-01

    To solve the problem of resonance during quasi-static calibration of high-g accelerometers, we deduce the relationship between the minimum excitation pulse width and the resonant frequency of the calibrated accelerometer according to the second-order mathematical model of the accelerometer, and improve the quasi-static calibration theory. We establish a quasi-static calibration testing system, which uses a gas gun to generate high-g acceleration signals, and apply a laser interferometer to reproduce the impact acceleration. These signals are used to drive the calibrated accelerometer. By comparing the excitation acceleration signal and the output responses of the calibrated accelerometer to the excitation signals, the impact sensitivity of the calibrated accelerometer is obtained. As indicated by the calibration test results, this calibration system produces excitation acceleration signals with a pulse width of less than 1000 μs, and realize the quasi-static calibration of high-g accelerometers with a resonant frequency above 20 kHz when the calibration error was 3%. PMID:28230743

  18. Ozone Correction for AM0 Calibrated Solar Cells for the Aircraft Method

    NASA Technical Reports Server (NTRS)

    Snyder, David B.; Scheiman, David A.; Jenkins, Phillip P.; Rieke, William J.; Blankenship, Kurt S.

    2002-01-01

    The aircraft solar cell calibration method has provided cells calibrated to space conditions for 37 years. However, it is susceptible to systematic errors due to ozone concentrations in the stratosphere. The present correction procedure applies a 1 percent increase to the measured I(sub SC) values. High band-gap cells are more sensitive to ozone absorbed wavelengths (0.4 to 0.8 microns) so it becomes important to reassess the correction technique. This paper evaluates the ozone correction to be 1+O3xFo, where O3 is the total ozone along the optical path, and Fo is 29.8 x 10(exp -6)/du for a Silicon solar cell, 42.6 x 10(exp -6)/du for a GaAs cell and 57.2 x 10(exp -6)/du for an InGaP cell. These correction factors work best to correct data points obtained during the flight rather than as a correction to the final result.

  19. A Single-Vector Force Calibration Method Featuring the Modern Design of Experiments

    NASA Technical Reports Server (NTRS)

    Parker, P. A.; Morton, M.; Draper, N.; Line, W.

    2001-01-01

    This paper proposes a new concept in force balance calibration. An overview of the state-of-the-art in force balance calibration is provided with emphasis on both the load application system and the experimental design philosophy. Limitations of current systems are detailed in the areas of data quality and productivity. A unique calibration loading system integrated with formal experimental design techniques has been developed and designated as the Single-Vector Balance Calibration System (SVS). This new concept addresses the limitations of current systems. The development of a quadratic and cubic calibration design is presented. Results from experimental testing are compared and contrasted with conventional calibration systems. Analyses of data are provided that demonstrate the feasibility of this concept and provide new insights into balance calibration.

  20. Quantifying uncertainty of determination by standard additions and serial dilutions methods taking into account standard uncertainties in both axes.

    PubMed

    Hyk, Wojciech; Stojek, Zbigniew

    2013-06-18

    The analytical expressions for the calculation of the standard uncertainty of the predictor variable either extrapolated or interpolated from a calibration line that takes into account uncertainties in both axes have been derived and successfully verified using the Monte Carlo modeling. These expressions are essential additions to the process of the analyte quantification realized with either the method of standard additions (SAM) or the method of serial dilutions (MSD). The latter one has been proposed as an alternative approach to the SAM procedure. In the MSD approach instead of the sequence of standard additions, the sequence of solvent additions to the spiked sample is performed. The comparison of the calculation results based on the expressions derived to their equivalents obtained from the Monte Carlo simulation, applied to real experimental data sets, confirmed that these expressions are valid in real analytical practice. The estimation of the standard uncertainty of the analyte concentration, quantified via either SAM or MSD or simply a calibration curve, is of great importance for the construction of the uncertainty budget of an analytical procedure. The correct estimation of the standard uncertainty of the analyte concentration is a key issue in the quality assurance in the instrumental analysis.

  1. System and method for high power diode based additive manufacturing

    DOEpatents

    El-Dasher, Bassem S.; Bayramian, Andrew; Demuth, James A.; Farmer, Joseph C.; Torres, Sharon G.

    2016-04-12

    A system is disclosed for performing an Additive Manufacturing (AM) fabrication process on a powdered material forming a substrate. The system may make use of a diode array for generating an optical signal sufficient to melt a powdered material of the substrate. A mask may be used for preventing a first predetermined portion of the optical signal from reaching the substrate, while allowing a second predetermined portion to reach the substrate. At least one processor may be used for controlling an output of the diode array.

  2. Application of Multivariate Linear and Nonlinear Calibration and Classification Methods in Drug Design.

    PubMed

    Abdolmaleki, Azizeh; Ghasemi, Jahan B; Shiri, Fereshteh; Pirhadi, Somayeh

    2015-01-01

    Data manipulation and maximum efficient extraction of useful information need a range of searching, modeling, mathematical, and statistical approaches. Hence, an adequate multivariate characterization is the first necessary step in investigation and the results are interpreted after multivariate analysis. Multivariate data analysis is capable of not only large dataset management but also interpret them surely and rapidly. Application of chemometrics and cheminformatics methods may be useful for design and discovery of new drug compounds. In this review, we present a variety of information sources on chemometrics, which we consider useful in different fields of drug design. This review describes exploratory analysis (PCA), classification and multivariate calibration (PCR, PLS) methods to data analysis. It summarizes the main facts of linear and nonlinear multivariate data analysis in drug discovery and provides an introduction to manipulation of data in this field. It handles the fundamental aspects of basic concepts of multivariate methods, principles of projections (PCA and PLS) and introduces the popular modeling and classification techniques. Enough theory behind these methods, more particularly concerning the chemometrics tools is included for those with little experience in multivariate data analysis techniques such as PCA, PLS, SIMCA, etc. We describe each method by avoiding unnecessary equations, and details of calculation algorithms. It provides a synopsis of the method followed by cases of applications in drug design (i.e., QSAR) and some of the features for each method.

  3. Addition and Subtraction. Mathematics-Methods Program Unit.

    ERIC Educational Resources Information Center

    LeBlanc, John F.; And Others

    This unit is 1 of 12 developed for the university classroom portion of the Mathematics-Methods Program (MMP), created by the Indiana University Mathematics Education Development Center (MEDC) as an innovative program for the mathematics training of prospective elementary school teachers (PSTs). Each unit is written in an activity format that…

  4. Sediment Core Extrusion Method at Millimeter Resolution Using a Calibrated, Threaded-rod

    PubMed Central

    Schwing, Patrick T.; Romero, Isabel C.; Larson, Rebekka A.; O'Malley, Bryan J.; Fridrik, Erika E.; Goddard, Ethan A.; Brooks, Gregg R.; Hastings, David W.; Rosenheim, Brad E.; Hollander, David J.; Grant, Guy; Mulhollan, Jim

    2016-01-01

    Aquatic sediment core subsampling is commonly performed at cm or half-cm resolution. Depending on the sedimentation rate and depositional environment, this resolution provides records at the annual to decadal scale, at best. An extrusion method, using a calibrated, threaded-rod is presented here, which allows for millimeter-scale subsampling of aquatic sediment cores of varying diameters. Millimeter scale subsampling allows for sub-annual to monthly analysis of the sedimentary record, an order of magnitude higher than typical sampling schemes. The extruder consists of a 2 m aluminum frame and base, two core tube clamps, a threaded-rod, and a 1 m piston. The sediment core is placed above the piston and clamped to the frame. An acrylic sampling collar is affixed to the upper 5 cm of the core tube and provides a platform from which to extract sub-samples. The piston is rotated around the threaded-rod at calibrated intervals and gently pushes the sediment out the top of the core tube. The sediment is then isolated into the sampling collar and placed into an appropriate sampling vessel (e.g., jar or bag). This method also preserves the unconsolidated samples (i.e., high pore water content) at the surface, providing a consistent sampling volume. This mm scale extrusion method was applied to cores collected in the northern Gulf of Mexico following the Deepwater Horizon submarine oil release. Evidence suggests that it is necessary to sample at the mm scale to fully characterize events that occur on the monthly time-scale for continental slope sediments. PMID:27585268

  5. Sediment Core Extrusion Method at Millimeter Resolution Using a Calibrated, Threaded-rod.

    PubMed

    Schwing, Patrick T; Romero, Isabel C; Larson, Rebekka A; O'Malley, Bryan J; Fridrik, Erika E; Goddard, Ethan A; Brooks, Gregg R; Hastings, David W; Rosenheim, Brad E; Hollander, David J; Grant, Guy; Mulhollan, Jim

    2016-08-17

    Aquatic sediment core subsampling is commonly performed at cm or half-cm resolution. Depending on the sedimentation rate and depositional environment, this resolution provides records at the annual to decadal scale, at best. An extrusion method, using a calibrated, threaded-rod is presented here, which allows for millimeter-scale subsampling of aquatic sediment cores of varying diameters. Millimeter scale subsampling allows for sub-annual to monthly analysis of the sedimentary record, an order of magnitude higher than typical sampling schemes. The extruder consists of a 2 m aluminum frame and base, two core tube clamps, a threaded-rod, and a 1 m piston. The sediment core is placed above the piston and clamped to the frame. An acrylic sampling collar is affixed to the upper 5 cm of the core tube and provides a platform from which to extract sub-samples. The piston is rotated around the threaded-rod at calibrated intervals and gently pushes the sediment out the top of the core tube. The sediment is then isolated into the sampling collar and placed into an appropriate sampling vessel (e.g., jar or bag). This method also preserves the unconsolidated samples (i.e., high pore water content) at the surface, providing a consistent sampling volume. This mm scale extrusion method was applied to cores collected in the northern Gulf of Mexico following the Deepwater Horizon submarine oil release. Evidence suggests that it is necessary to sample at the mm scale to fully characterize events that occur on the monthly time-scale for continental slope sediments.

  6. Measurement of cardiac output by earpiece dye-dilution method with automatic calibration of dye concentration.

    PubMed

    Katori, R; Hayashi, T; Kanamasa, K; Ishikawa, K

    1977-05-01

    A non-invasive method for measuring cardiac output by an earpiece dye densitometer was proposed. The densitometer is dichromatic and has an air capsule to make subject's ear bloodless by inflation, so that it can calibrate indocyanine green dye concentration without blood sampling. Duplicate measurements of cardiac output showed a good agreement in 40 cases (r = 0.97, standard deviation (S.D.) = 8.9%), which was comparable to the result of the cuvette method (r = 0.98, S.D. = 8.1%). Simultaneous measurements of cardiac output by the earpiece method (x) and the standard cuvette method (y) revealed a good agreement (r = 0.91, y = 0.96x+0.34, S.D. = 16.4%) in 52 measurements of 25 cases. A similarly good correlation was obtained between the two methods during ergometer exercise in supine position in 5 cases. These suggest that this earpiece dye-dilution method is reliable for cardiac output measurement and advantageous for clinical use because of non-invasive technique.

  7. A New Method for Calibrating Perceptual Salience across Dimensions in Infants: The Case of Color vs. Luminance

    ERIC Educational Resources Information Center

    Kaldy, Zsuzsa; Blaser, Erik A.; Leslie, Alan M.

    2006-01-01

    We report a new method for calibrating differences in perceptual salience across feature dimensions, in infants. The problem of inter-dimensional salience arises in many areas of infant studies, but a general method for addressing the problem has not previously been described. Our method is based on a preferential looking paradigm, adapted to…

  8. Development of a New Calibration Method for an Ambient Ion Monitor Ion Chromatograph (AIM-IC)

    NASA Astrophysics Data System (ADS)

    Markovic, M.; Vandenboer, T.; Murphy, J. G.

    2009-05-01

    Fine atmospheric aerosols play an important role in the atmosphere as they alter the radiative balance of the Earth through direct and indirect climate effects, reduce visibility, participate in acid rain formation and affect human health. The motivation for chemically and temporally resolved measurements of fine aerosol composition has lead to the development of the Ambient Ion Monitor Ion Chromatograph (AIM-IC) system by Dionex/URG. This instrument is capable of simultaneously monitoring fine aerosols (<2.5μm) and associated precursor gases on a nearly continuous basis with a time resolution of 1 hour. The instrument utilizes a parallel-plate wet denuder with a constantly regenerated surface for collection of gases and a particle condensation chamber for the collection of aerosols. AIM-IC is capable of monitoring HCl(g), HONO(g), HNO3(g), SO2(g), NH3(g), Cl-, NO2-, NO3-, SO42-, NH4+ , and some water soluble organic acids and amines. Standard calibration of the AIM-IC is carried out by injecting a series of mixed standards directly onto the ion chromatographs, bypassing the sampling component of the instrument. This results in calculated detection limits on the order of 10-200 pptv for gases and 10-500 of ng/m3 for individual particle constituents when collecting at 3 L/min for 55 minutes. In this work, we present a new method for the calibration of the AIM-IC for both gas and particle collection that enables us to evaluate the entire system from size-selection to detection. This external calibration method is assessed for the gases HNO3(g), SO2(g), and NH3(g), and for particles containing (NH4)2SO4, NH4NO3, and Na2SO4. Quantitative collection of SO2 is found to require careful optimization of the H2O2 concentration of the denuder liquid, while the replacement of a cyclone with an impactor improves the sampling efficiency of NH3 and HNO3.

  9. Simultaneous spectrophotometric determination of chromium(VI) and iron (III) by H-point standard addition method

    NASA Astrophysics Data System (ADS)

    Larionova, E. V.; Bulygina, K. A.

    2016-02-01

    In this work the possibility of simultaneous spectrophotometric determination of chromium (VI) and iron (III) in alloys with help of the mixed organic reagent (diphenylcarbazide and 1,10-phenanthroline) is studied. We have applied H-point standard addition method to determine concentrations of chromium (VI) and iron (III) from the mixture. The pure signals of complexes of chromium (VI) with diphenylcarbazide and iron (III) with the 1,10-phenanthroline and their calibration plots are previously carried out. We established the possibility of simultaneous determination of chromium (VI) and iron (III) in the different concentration ranges by H-point standard addition method. Correctness of determination of concentration by means of the offered technique is proved by "added-found" method for a series of mixtures with different ratios of concentration of chromium (VI) and iron (III). It is founded that the error of determination of concentration doesn't exceed 33%.

  10. Methods for Optical Calibration of the BigBite Hadron Spectrometer

    SciTech Connect

    M. Mihovilovic, K. Allada, B.D. Anderson, J.R.M. Annand, T. Averett, A. Camsonne, R.W. Chan, J.-P. Chen, K. Chirapatpimol, C.W. de Jager, S. Gilad, D.J. Hamilton, J.-O. Hansen, D.W. Higinbotham, J. Juang, X. Jiang, G. Jin, W. Korsch, J.J. LeRose, R.A. Lindgren, N. Liyanage, E. Long, R. Michaels, B. Moffit, P. Monaghan, V. Nelyubin, B.E. Norum, E. Piasetzky, X. Qian, Y. Qiang, S. Riordan, G. Ron, G. Rosner, B. Sawatzky, M. Shabestari, A. Shahinyan, R. Shneor, S. Sirca, R. Subedi, V. Sulkosky, J.W. Watson, B. Wojtsekhowski, Y.-W. Zhang

    2012-09-01

    The techniques for optical calibration of Jefferson Lab's large-acceptance magnetic hadron spectrometer, BigBite, have been examined. The most consistent and stable results were obtained by using a method based on singular value decomposition. In spite of the complexity of the optics, the particles positions and momenta at the target have been precisely reconstructed from the coordinates measured in the detectors by means of a single back-tracing matrix. The technique is applicable to any similar magnetic spectrometer and any particle type. For 0.55 GeV/c protons, we have established the vertex resolution of 1.2 cm, angular resolutions of 7 mrad and 16 mrad (in-plane and out-of-plane, respectively), and a relative momentum resolution of 1.6%.

  11. A Simple Permittivity Calibration Method for GPR-Based Road Pavement Measurements

    NASA Astrophysics Data System (ADS)

    Eskelinen, Pekka

    2016-09-01

    Cylindrical resonator principle can be used in GPR asphalt quality measurement calibration. This method relies on ordinary drill core samples that are regularly taken from measured road sections, but now only analyzed for dimensions, density and sometimes chemically. If such a drill sample is covered with proper conductive surfaces, a cylindrical cavity resonator is formed. The baseline of the GPR permittivity recordings can so be found by measuring the resonance behaviour of this covered sample, which can later still be used for those traditional analyses. A clear benefit is the resonator's 1-2 GHz frequency range which equals that of common commercial GPR systems. Example results and reference readings from known dielectric material are shown. The obtained uncertainty in this case study is 0.02 units of permittivity, when measuring the same sample repeatedly.

  12. Spectrum reconstruction method based on the detector response model calibrated by x-ray fluorescence.

    PubMed

    Li, Ruizhe; Li, Liang; Chen, Zhiqiang

    2017-02-07

    Accurate estimation of distortion-free spectra is important but difficult in various applications, especially for spectral computed tomography. Two key problems must be solved to reconstruct the incident spectrum. One is the acquisition of the detector energy response. It can be calculated by Monte Carlo simulation, which requires detailed modeling of the detector system and a high computational power. It can also be acquired by establishing a parametric response model and be calibrated using monochromatic x-ray sources, such as synchrotron sources or radioactive isotopes. However, these monochromatic sources are difficult to obtain. Inspired by x-ray fluorescence (XRF) spectrum modeling, we propose a feasible method to obtain the detector energy response based on an optimized parametric model for CdZnTe or CdTe detectors. The other key problem is the reconstruction of the incident spectrum with the detector response. Directly obtaining an accurate solution from noisy data is difficult because the reconstruction problem is severely ill-posed. Different from the existing spectrum stripping method, a maximum likelihood-expectation maximization iterative algorithm is developed based on the Poisson noise model of the system. Simulation and experiment results show that our method is effective for spectrum reconstruction and markedly increases the accuracy of XRF spectra compared with the spectrum stripping method. The applicability of the proposed method is discussed, and promising results are presented.

  13. Spectrum reconstruction method based on the detector response model calibrated by x-ray fluorescence

    NASA Astrophysics Data System (ADS)

    Li, Ruizhe; Li, Liang; Chen, Zhiqiang

    2017-02-01

    Accurate estimation of distortion-free spectra is important but difficult in various applications, especially for spectral computed tomography. Two key problems must be solved to reconstruct the incident spectrum. One is the acquisition of the detector energy response. It can be calculated by Monte Carlo simulation, which requires detailed modeling of the detector system and a high computational power. It can also be acquired by establishing a parametric response model and be calibrated using monochromatic x-ray sources, such as synchrotron sources or radioactive isotopes. However, these monochromatic sources are difficult to obtain. Inspired by x-ray fluorescence (XRF) spectrum modeling, we propose a feasible method to obtain the detector energy response based on an optimized parametric model for CdZnTe or CdTe detectors. The other key problem is the reconstruction of the incident spectrum with the detector response. Directly obtaining an accurate solution from noisy data is difficult because the reconstruction problem is severely ill-posed. Different from the existing spectrum stripping method, a maximum likelihood-expectation maximization iterative algorithm is developed based on the Poisson noise model of the system. Simulation and experiment results show that our method is effective for spectrum reconstruction and markedly increases the accuracy of XRF spectra compared with the spectrum stripping method. The applicability of the proposed method is discussed, and promising results are presented.

  14. Monitoring Method of Cutting Force by Using Additional Spindle Sensors

    NASA Astrophysics Data System (ADS)

    Sarhan, Ahmed Aly Diaa; Matsubara, Atsushi; Sugihara, Motoyuki; Saraie, Hidenori; Ibaraki, Soichi; Kakino, Yoshiaki

    This paper describes a monitoring method of cutting forces for end milling process by using displacement sensors. Four eddy-current displacement sensors are installed on the spindle housing of a machining center so that they can detect the radial motion of the rotating spindle. Thermocouples are also attached to the spindle structure in order to examine the thermal effect in the displacement sensing. The change in the spindle stiffness due to the spindle temperature and the speed is investigated as well. Finally, the estimation performance of cutting forces using the spindle displacement sensors is experimentally investigated by machining tests on carbon steel in end milling operations under different cutting conditions. It is found that the monitoring errors are attributable to the thermal displacement of the spindle, the time lag of the sensing system, and the modeling error of the spindle stiffness. It is also shown that the root mean square errors between estimated and measured amplitudes of cutting forces are reduced to be less than 20N with proper selection of the linear stiffness.

  15. Ultrasonic self-calibrated method applied to monitoring of sol-gel transition.

    PubMed

    Robin, Guillaume; Vander Meulen, François; Wilkie-Chancellier, Nicolas; Martinez, Loïc; Haumesser, Lionel; Fortineau, Jérôme; Griesmar, Pascal; Lethiecq, Marc; Feuillard, Guy

    2012-07-01

    In many industrial processes where online control is necessary such as in the food industry, the real time monitoring of visco-elastic properties is essential to ensure the quantity of production. Acoustic methods have shown that reliable properties could be obtained from measurements of velocity and attenuation. This paper proposes a simple, real time ultrasound method for monitoring linear medium properties (phase velocity and attenuation) that vary in time. The method is based on a pulse echo measurement and is self-calibrated. Results on a silica gel are reported and the importance of taking into account the changes of the mechanical loading on the front face of the transducer will be shown. This is done through a modification of the emission and reception transfer parameters. The simultaneous measurement of the input and output currents and voltages enables these parameters to be calculated during the reaction. The variations of the transfer parameters are in the order of 6% and predominate other effects. The evolution of the ultrasonic longitudinal wave phase velocity and attenuation as a function of time allows the characteristic times of the chemical reaction to be determined. The results are well correlated with the gelation time measured by rheological method at low frequency.

  16. Multi-Axis Accelerometer Calibration System

    NASA Technical Reports Server (NTRS)

    Finley, Tom; Parker, Peter

    2010-01-01

    A low-cost, portable, and simplified system has been developed that is suitable for in-situ calibration and/or evaluation of multi-axis inertial measurement instruments. This system overcomes facility restrictions and maintains or improves the calibration quality for users of accelerometer-based instruments with applications in avionics, experimental wind tunnel research, and force balance calibration applications. The apparatus quickly and easily positions a multi-axis accelerometer system into a precisely known orientation suitable for in-situ quality checks and calibration. In addition, the system incorporates powerful and sophisticated statistical methods, known as response surface methodology and statistical quality control. These methods improve calibration quality, reduce calibration time, and allow for increased calibration frequency, which enables the monitoring of instrument stability over time.

  17. An innovative method for coordinate measuring machine one-dimensional self-calibration with simplified experimental process.

    PubMed

    Fang, Cheng; Butler, David Lee

    2013-05-01

    In this paper, an innovative method for CMM (Coordinate Measuring Machine) self-calibration is proposed. In contrast to conventional CMM calibration that relies heavily on a high precision reference standard such as a laser interferometer, the proposed calibration method is based on a low-cost artefact which is fabricated with commercially available precision ball bearings. By optimizing the mathematical model and rearranging the data sampling positions, the experimental process and data analysis can be simplified. In mathematical expression, the samples can be minimized by eliminating the redundant equations among those configured by the experimental data array. The section lengths of the artefact are measured at arranged positions, with which an equation set can be configured to determine the measurement errors at the corresponding positions. With the proposed method, the equation set is short of one equation, which can be supplemented by either measuring the total length of the artefact with a higher-precision CMM or calibrating the single point error at the extreme position with a laser interferometer. In this paper, the latter is selected. With spline interpolation, the error compensation curve can be determined. To verify the proposed method, a simple calibration system was set up on a commercial CMM. Experimental results showed that with the error compensation curve uncertainty of the measurement can be reduced to 50%.

  18. Instrumentation and calibration methods for the multichannel measurement of phase and amplitude in optical tomography

    SciTech Connect

    Nissilae, Ilkka; Noponen, Tommi; Kotilahti, Kalle; Katila, Toivo; Lipiaeinen, Lauri; Tarvainen, Tanja; Schweiger, Martin; Arridge, Simon

    2005-04-01

    In this article, we describe the multichannel implementation of an intensity modulated optical tomography system developed at Helsinki University of Technology. The system has two time-multiplexed wavelengths, 16 time-multiplexed source fibers and 16 parallel detection channels. The gain of the photomultiplier tubes (PMTs) is individually adjusted during the measurement sequence to increase the dynamic range of the system by 10{sup 4}. The PMT used has a high quantum efficiency in the near infrared (8% at 800 nm), a fast settling time, and low hysteresis. The gain of the PMT is set so that the dc anode current is below 80 nA, which allows the measurement of phase independently of the intensity. The system allows measurements of amplitude at detected intensities down to 1 fW, which is sufficient for transmittance measurements of the female breast, the forearm, and the brain of early pre-term infants. The mean repeatability of phase and the logarithm of amplitude (ln A) at 100 MHz were found to be 0.08 deg. and 0.004, respectively, in a measurement of a 7 cm phantom with an imaging time of 5 s per source and source optical power of 8 mW. We describe a three-step method of calibrating the phase and amplitude measurements so that the absolute absorption and scatter in tissue may be measured. A phantom with two small cylindrical targets and a second phantom with three rods are measured and reconstructions made from the calibrated data are shown and compared with reconstructions from simulated data.

  19. Technique for Radiometer and Antenna Array Calibration with Two Antenna Noise Diodes

    NASA Technical Reports Server (NTRS)

    Srinivasan, Karthik; Limaye, Ashutosh; Laymon, Charles; Meyer, Paul

    2011-01-01

    This paper presents a new technique to calibrate a microwave radiometer and phased array antenna system. This calibration technique uses a radiated noise source in addition to an injected noise sources for calibration. The plane of reference for this calibration technique is the face of the antenna and therefore can effectively calibration the gain fluctuations in the active phased array antennas. This paper gives the mathematical formulation for the technique and discusses the improvements brought by the method over the existing calibration techniques.

  20. A new method for calibrating a boron isotope paleo-pH proxy using massive Porites corals

    NASA Astrophysics Data System (ADS)

    Kubota, Kaoru; Yokoyama, Yusuke; Ishikawa, Tsuyoshi; Suzuki, Atsushi

    2015-09-01

    The boron isotope ratio (δ11B) of marine biogenic carbonates can reconstruct pH and pCO2 of seawater, and potentially CO2 concentration in the atmosphere. To date, δ11B-pHSW calibration has been proposed via culturing experiments, where calcifying organisms are cultured under artificially acidified seawater. However, in scleractinian corals, reconstructed pH values using culture-based calibrations do not agree well with actual observations of seawater CO2 chemistry. Thus, another approach is needed to establish a more reliable calibration method. In this study, we established field-based calibrations for Chichijima and Tahiti, both located in subtropical gyres where surface seawater is close to CO2 equilibrium. We suggest a new approach to calibration of δ11B-pH in which the long-term δ11B variation of massive Porites corals is compared with the decreasing pH trend (i.e., ocean acidification) that has occurred since the Industrial Revolution. This calibration will offer a new avenue for studying seawater CO2 chemistry using coral δ11B in diverse settings, such as upwelling regions, coral reefs, and coastal areas.

  1. A Comparative Study of Online Pretest Item Calibration/Scaling Methods in CAT. ACT Research Report Series.

    ERIC Educational Resources Information Center

    Ban, Jae-Chun; Hanson, Bradley A.; Wang, Tianyou; Yi, Qing; Harris, Deborah J.

    The purpose of this study was to compare and evaluate five online pretest item calibration/scaling methods in computerized adaptive testing (CAT): (1) the marginal maximum likelihood estimate with one-EM cycle (OEM); (2) the marginal maximum likelihood estimate with multiple EM cycles (MEM); (3) Stocking's Method A (M. Stocking, 1988); (4)…

  2. Array error calibration methods in downward-looking linear-array three-dimensional synthetic aperture radar

    NASA Astrophysics Data System (ADS)

    Tan, Weixian; Huang, Pingping; Han, Kuoye; Liu, Qi; Peng, Xueming

    2016-04-01

    In order to achieve high-precision three-dimensional (3-D) imaging with an airborne downward-looking linear-array 3-D synthetic aperture radar (LA-3D-SAR), a uniform virtual antenna array can be obtained by aperture synthesis of the cross-track sparse multiple-input-multiple-output array. However, the actual 3-D imaging quality is unavoidably degraded by array errors such as the multichannel amplitude-phase errors due to the nonideal antenna characteristics, and the virtual element position errors due to vibrations and motion measurement deviations. We investigate the effects of these errors on the forms and the degrees of image quality degradation and consider the use of corresponding calibration methods to eliminate the effects of errors. For the multichannel amplitude-phase errors, the target response is subject to an integrated sidelobe level increase introduced by the phase error, which can be calibrated based on external (parallel or point target) calibrators, as proposed in the paper. For the virtual element position errors, they mainly the result of contrast degradation and noise in the image along the cross-track direction and have little impact on the range and along-track directions. The imaging performance is more sensitive to the error component in the height direction as compared to other components, the precision requirement of which should be established as the calibration reference. A calibration method based on time-divided active calibrators is proposed to estimate and correct the virtual element position errors. Both numerical simulations and real data experiments have shown the validity of the analyses as well as the effectiveness of the proposed calibration methods.

  3. A novel INS and Doppler sensors calibration method for long range underwater vehicle navigation.

    PubMed

    Tang, Kanghua; Wang, Jinling; Li, Wanli; Wu, Wenqi

    2013-10-28

    Since the drifts of Inertial Navigation System (INS) solutions are inevitable and also grow over time, a Doppler Velocity Log (DVL) is used to aid the INS to restrain its error growth. Therefore, INS/DVL integration is a common approach for Autonomous Underwater Vehicle (AUV) navigation. The parameters including the scale factor of DVL and misalignments between INS and DVL are key factors which limit the accuracy of the INS/DVL integration. In this paper, a novel parameter calibration method is proposed. An iterative implementation of the method is designed to reduce the error caused by INS initial alignment. Furthermore, a simplified INS/DVL integration scheme is employed. The proposed method is evaluated with both river trial and sea trial data sets. Using 0.03°/h(1σ) ring laser gyroscopes, 5 × 10-5 g(1σ) quartz accelerometers and DVL with accuracy 0.5% V ± 0.5 cm/s, INS/DVL integrated navigation can reach an accuracy of about 1‰ of distance travelled (CEP) in a river trial and 2‰ of distance travelled (CEP) in a sea trial.

  4. Self-Calibrating and Remote Programmable Signal Conditioning Amplifier System and Method

    NASA Technical Reports Server (NTRS)

    Medelius, Pedro J. (Inventor); Hallberg, Carl G. (Inventor); Simpson, Howard J., III (Inventor); Thayer, Stephen W. (Inventor)

    1998-01-01

    A self-calibrating, remote programmable signal conditioning amplifier system employs information read from a memory attached to a measurement transducer for automatic calibration. The signal conditioning amplifier is self-calibrated on a continuous basis through use of a dual input path arrangement, with each path containing a multiplexer and a programmable amplifier. A digital signal processor controls operation of the system such that a transducer signal is applied to one of the input paths, while one or more calibration signals are applied to the second input path. Once the second path is calibrated, the digital signal processor switches the transducer signal to the second path. and then calibrates the first path. This process is continually repeated so that each path is calibrated on an essentially continuous basis. Dual output paths are also employed which are calibrated in the same manner. The digital signal processor also allows the implementation of a variety of digital filters which are either programmed into the system or downloaded by an operator, and performs up to eighth order linearization.

  5. An in situ hand calibration method using a pseudo-observation scheme for low-end inertial measurement units

    NASA Astrophysics Data System (ADS)

    Li, You; Niu, Xiaoji; Zhang, Quan; Zhang, Hongping; Shi, Chuang

    2012-10-01

    MEMS chips have become ideal candidates for various applications since they are small sized, light weight, have low power consumption and are extremely low cost and reliable. However, the performance of MEMS sensors, especially their biases and scale factors, is highly dependent on environmental conditions such as temperature. Thus a quick and convenient calibration is needed to be conducted by users in field without any external equipment or any expert knowledge of calibration. A novel and efficient in situ hand calibration method is presented to meet these demands in this paper. The algorithm of the proposed calibration method makes use of the navigation algorithm of the loosely-coupled GPS/INS integrated systems, but replaces the GPS observations with a kind of pseudo-observations, which can be stated as follows: if an inertial measurement unit (IMU) was rotating approximately around its measurement center, the range of its position and its linear velocity both would be within a limited scope. Using a Kalman filtering algorithm, the biases and scale factors of both accelerometer triad and gyroscope triad can be calibrated together within a short period (about 30 s), requiring only motions by hands. Real test results show that the proposed method is suitable for most consumer grade MEMS IMUs due to its zero cost, easy operation and sufficient accuracy.

  6. Calibration of measurement sensitivities of multiple micro-cantilever dynamic modes in atomic force microscopy using a contact detection method

    SciTech Connect

    Liu Zhen; Jeong, Younkoo; Menq, Chia-Hsiang

    2013-02-15

    An accurate experimental method is proposed for on-spot calibration of the measurement sensitivities of multiple micro-cantilever dynamic modes in atomic force microscopy. One of the key techniques devised for this method is a reliable contact detection mechanism that detects the tip-surface contact instantly. At the contact instant, the oscillation amplitude of the tip deflection, converted to that of the deflection signal in laser reading through the measurement sensitivity, exactly equals to the distance between the sample surface and the cantilever base position. Therefore, the proposed method utilizes the recorded oscillation amplitude of the deflection signal and the base position of the cantilever at the contact instant for the measurement sensitivity calibration. Experimental apparatus along with various signal processing and control modules was realized to enable automatic and rapid acquisition of multiple sets of data, with which the calibration of a single dynamic mode could be completed in less than 1 s to suppress the effect of thermal drift and measurement noise. Calibration of the measurement sensitivities of the first and second dynamic modes of three micro-cantilevers having distinct geometries was successfully demonstrated. The dependence of the measurement sensitivity on laser spot location was also experimentally investigated. Finally, an experiment was performed to validate the calibrated measurement sensitivity of the second dynamic mode of a micro-cantilever.

  7. Calibration Method for ML Estimation of 3D Interaction Position in a Thick Gamma-Ray Detector

    PubMed Central

    Hunter, William C. J.; Barrett, Harrison H.; Furenlid, Lars R.

    2010-01-01

    High-energy (> 100 keV) photon detectors are often made thick relative to their lateral resolution in order to improve their photon-detection efficiency. To avoid issues of parallax and increased signal variance that result from random interaction depth, we must determine the 3D interaction position in the imaging detector. With this goal in mind, we examine a method of calibrating response statistics of a thick-detector gamma camera to produce a maximum-likelihood estimate of 3D interaction position. We parameterize the mean detector response as a function of 3D position, and we estimate these parameters by maximizing their likelihood given prior knowledge of the pathlength distribution and a complete list of camera signals for an ensemble of gamma-ray interactions. Furthermore, we describe an iterative method for removing multiple-interaction events from our calibration data and for refining our calibration of the mean detector response to single interactions. We demonstrate this calibration method with simulated gamma-camera data. We then show that the resulting calibration is accurate and can be used to produce unbiased estimates of 3D interaction position. PMID:20191099

  8. Calibration of measurement sensitivities of multiple micro-cantilever dynamic modes in atomic force microscopy using a contact detection method.

    PubMed

    Liu, Zhen; Jeong, Younkoo; Menq, Chia-Hsiang

    2013-02-01

    An accurate experimental method is proposed for on-spot calibration of the measurement sensitivities of multiple micro-cantilever dynamic modes in atomic force microscopy. One of the key techniques devised for this method is a reliable contact detection mechanism that detects the tip-surface contact instantly. At the contact instant, the oscillation amplitude of the tip deflection, converted to that of the deflection signal in laser reading through the measurement sensitivity, exactly equals to the distance between the sample surface and the cantilever base position. Therefore, the proposed method utilizes the recorded oscillation amplitude of the deflection signal and the base position of the cantilever at the contact instant for the measurement sensitivity calibration. Experimental apparatus along with various signal processing and control modules was realized to enable automatic and rapid acquisition of multiple sets of data, with which the calibration of a single dynamic mode could be completed in less than 1 s to suppress the effect of thermal drift and measurement noise. Calibration of the measurement sensitivities of the first and second dynamic modes of three micro-cantilevers having distinct geometries was successfully demonstrated. The dependence of the measurement sensitivity on laser spot location was also experimentally investigated. Finally, an experiment was performed to validate the calibrated measurement sensitivity of the second dynamic mode of a micro-cantilever.

  9. Towards monolithic scintillator based TOF-PET systems: practical methods for detector calibration and operation

    NASA Astrophysics Data System (ADS)

    Borghi, Giacomo; Tabacchini, Valerio; Schaart, Dennis R.

    2016-07-01

    Gamma-ray detectors based on thick monolithic scintillator crystals can achieve spatial resolutions  <2 mm full-width-at-half-maximum (FWHM) and coincidence resolving times (CRTs) better than 200 ps FWHM. Moreover, they provide high sensitivity and depth-of-interaction (DOI) information. While these are excellent characteristics for clinical time-of-flight (TOF) positron emission tomography (PET), the application of monolithic scintillators has so far been hampered by the lengthy and complex procedures needed for position- and time-of-interaction estimation. Here, the algorithms previously developed in our group are revised to make the calibration and operation of a large number of monolithic scintillator detectors in a TOF-PET system practical. In particular, the k-nearest neighbor (k-NN) classification method for x,y-position estimation is accelerated with an algorithm that quickly preselects only the most useful reference events, reducing the computation time for position estimation by a factor of ~200 compared to the previously published k-NN 1D method. Also, the procedures for estimating the DOI and time of interaction are revised to enable full detector calibration by means of fan-beam or flood irradiations only. Moreover, a new technique is presented to allow the use of events in which some of the photosensor pixel values and/or timestamps are missing (e.g. due to dead time), so as to further increase system sensitivity. The accelerated methods were tested on a monolithic scintillator detector specifically developed for clinical PET applications, consisting of a 32 mm  ×  32 mm  ×  22 mm LYSO : Ce crystal coupled to a digital photon counter (DPC) array. This resulted in a spatial resolution of 1.7 mm FWHM, an average DOI resolution of 3.7 mm FWHM, and a CRT of 214 ps. Moreover, the possibility of using events missing the information of up to 16 out of 64 photosensor pixels is shown. This results in only a small

  10. Application of an Artificial Intelligence Method for Velocity Calibration and Events Location in Microseismic Monitoring

    NASA Astrophysics Data System (ADS)

    Yang, Y.; Chen, X.

    2013-12-01

    Good quality hydraulic fracture maps are heavily dependent upon the best possible velocity structure. Particle Swarm Optimization inversion scheme, an artificial intelligence technique for velocity calibration and events location could serve as a viable option, able to produce high quality data. Using perforation data to recalibrate the 1D isotropic velocity model derived from dipole sonic logs (or even without them), we are able to get the initial velocity model used for consequential events location. Velocity parameters can be inverted, as well as the thickness of the layer, through an iterative procedure. Performing inversion without integrating available data is unlikely to produce reliable results; especially if there are only one perforation shot and a single poor-layer-covered array along with low signal/noise ratio signal. The inversion method was validated via simulations and compared to the Fast Simulated Annealing approach and the Conjugate Gradient method. Further velocity model refinement can be accomplished while calculating events location during the iterative procedure minimizing the residuals from both sides. This artificial intelligence technique also displays promising application to the joint inversion of large-scale seismic activities data.

  11. a Novel Approach to Camera Calibration Method for Smart Phones Under Road Environment

    NASA Astrophysics Data System (ADS)

    Lee, Bijun; Zhou, Jian; Ye, Maosheng; Guo, Yuan

    2016-06-01

    Monocular vision-based lane departure warning system has been increasingly used in advanced driver assistance systems (ADAS). By the use of the lane mark detection and identification, we proposed an automatic and efficient camera calibration method for smart phones. At first, we can detect the lane marker feature in a perspective space and calculate edges of lane markers in image sequences. Second, because of the width of lane marker and road lane is fixed under the standard structural road environment, we can automatically build a transformation matrix between perspective space and 3D space and get a local map in vehicle coordinate system. In order to verify the validity of this method, we installed a smart phone in the `Tuzhi' self-driving car of Wuhan University and recorded more than 100km image data on the road in Wuhan. According to the result, we can calculate the positions of lane markers which are accurate enough for the self-driving car to run smoothly on the road.

  12. [A calibrated method for blood pressure measurement based on volume pulse wave].

    PubMed

    Youde, Ding; Qinkai, Deng; Feixue, Liang; Jinseng, Guo

    2010-01-01

    Physiology parameters measurement based on volume pulse wave is suitable for the monitoring blood pressure continuously. This paper described that the systolic blood pressure (SBP) and diastolic blood pressure (DBP) can be calibrated by measuring the pulse propagation time, just on one point of finger tip. The volume pulse wave was acquired by lighting the red and infrared LED alternately, and after signal processing, an accelerated pulse wave was obtained. Then by measuring the pulse wave propagation time between the progressive wave and reflected wave, we can find the relationship of the time and the blood pressure, and establish the related systolic blood pressure measurement equation. At the same time, based on the relationship between alternating current and direct current components in the volume pulse waveforms and through regression analysising, the relevant diastolic blood pressure measurement equation can be established. 33 clinical experimentation cases have been worked by dividing them into two groups: training group (18 cases) and control group (15 cases), by comparing with the measuring results of the OMRON electronic sphygmomanometer. The results indicated that the two methods had good coherence. The measurement described is simple and reliable, and may be served as a new method for noninvasively and continuously measurement of blood pressure.

  13. Performance of Higher Order Campbell methods, Part II: calibration and experimental application

    NASA Astrophysics Data System (ADS)

    Elter, Zs.; de Izarra, G.; Filliatre, P.; Jammes, C.; Pázsit, I.

    2016-11-01

    Applying Higher Order Campbelling methods in neutron flux monitoring with fission chambers is advantageous due to their capabilities to suppress the impact of unwanted noises and signal contributions (such as gamma radiation). This work aims to verify through experimental results that the basic assumptions behind the Higher Order Campelling methods are valid in critical reactors. The experiments, reported in this work, were performed at the MINERVE reactor in Cadarache. It is shown that the calibration of a fission chamber and the associated electronic system is possible in higher order mode. With the use of unbiased cumulant estimators and with digital processing, it is shown that over a wide count rate range, accurate count rate estimation can be achieved based on signal samples of a few ms, which is a significant progress compared to similar experimental results in the literature. The difference between the count rate estimated by pulse counting and by the Higher Order Campelling is less than 4%. The work also investigates the possibility of monitoring transient events. For this purpose, a control rod drop event was followed in Higher Order Campbelling mode.

  14. Method for calibrating a Fourier transform ion cyclotron resonance mass spectrometer

    DOEpatents

    Smith, Richard D.; Masselon, Christophe D.; Tolmachev, Aleksey

    2003-08-19

    A method for improving the calibration of a Fourier transform ion cyclotron resonance mass spectrometer wherein the frequency spectrum of a sample has been measured and the frequency (f) and intensity (I) of at least three species having known mass to charge (m/z) ratios and one specie having an unknown (m/z) ratio have been identified. The method uses the known (m/z) ratios, frequencies, and intensities at least three species to calculate coefficients A, B, and C, wherein the mass to charge ratio of a least one of the three species (m/z).sub.i is equal to ##EQU1## wherein f.sub.i is the detected frequency of the specie, G(I.sub.i) is a predetermined function of the intensity of the species, and Q is a predetermined exponent. Using the calculated values for A, B, and C, the mass to charge ratio of the unknown specie (m/z).sub.ii is calculated as the sum of ##EQU2## wherein f.sub.ii is the measured frequency of the unknown specie, and (I.sub.ii) is the measured intensity of the unknown specie.

  15. Satellite-Sensor Calibration Verification Using the Cloud-Shadow Method

    NASA Technical Reports Server (NTRS)

    Reinersman, P.; Carder, K. L.; Chen, F. R.

    1995-01-01

    An atmospheric-correction method which uses cloud-shaded pixels together with pixels in a neighboring region of similar optical properties is described. This cloud-shadow method uses the difference between the total radiance values observed at the sensor for these two regions, thus removing the nearly identical atmospheric radiance contributions to the two signals (e.g. path radiance and Fresnel-reflected skylight). What remains is largely due to solar photons backscattered from beneath the sea to dominate the residual signal. Normalization by the direct solar irradiance reaching the sea surface and correction for some second-order effects provides the remote-sensing reflectance of the ocean at the location of the neighbor region, providing a known 'ground target' spectrum for use in testing the calibration of the sensor. A similar approach may be useful for land targets if horizontal homogeneity of scene reflectance exists about the shadow. Monte Carlo calculations have been used to correct for adjacency effects and to estimate the differences in the skylight reaching the shadowed and neighbor pixels.

  16. A Novel Non-Intrusive Method to Resolve the Thermal-Dome-Effect of Pyranometers: Radiometric Calibration and Implications

    NASA Technical Reports Server (NTRS)

    Ji, Qiang; Tsay, Si-Chee; Lau, K. M.; Hansell, R. A.; Butler, J. J.; Cooper, J. W.

    2011-01-01

    Traditionally the calibration equation for pyranometers assumes that the measured solar irradiance is solely proportional to the thermopile's output voltage; therefore only a single calibration factor is derived. This causes additional measurement uncertainties because it does not capture sufficient information to correctly account for a pyranometer's thermal effect. In our updated calibration equation, temperatures from the pyranometer's dome and case are incorporated to describe the instrument's thermal behavior, and a new set of calibration constants are determined, thereby reducing measurement uncertainties. In this paper, we demonstrate why a pyranometer's uncertainty using the traditional calibration equation is always larger than a-few-percent, but with the new approach can become much less than 1% after the thermal issue is resolved. The highlighted calibration results are based on NIST-traceable light sources under controlled laboratory conditions. The significance of the new approach lends itself to not only avoiding the uncertainty caused by a pyranometer's thermal effect but also the opportunity to better isolate and characterize other instrumental artifacts, such as angular response and non-linearity of the thermopile, to further reduce additional uncertainties. We also discuss some of the implications, including an example of how the thermal issue can potentially impact climate studies by evaluating aerosol's direct-radiative effect using field measurements with and without considering the pyranometer's thermal effect. The results of radiative transfer model simulation show that a pyranometer's thermal effect on solar irradiance measurements at the surface can be translated into a significant alteration of the calculated distribution of solar energy inside the column atmosphere.

  17. Calibration of sound calibrators: an overview

    NASA Astrophysics Data System (ADS)

    Milhomem, T. A. B.; Soares, Z. M. D.

    2016-07-01

    This paper presents an overview of calibration of sound calibrators. Initially, traditional calibration methods are presented. Following, the international standard IEC 60942 is discussed emphasizing parameters, target measurement uncertainty and criteria for conformance to the requirements of the standard. Last, Regional Metrology Organizations comparisons are summarized.

  18. Methods of travel-time residual declustering for the knowledge base calibration and integration tool (KBCIT)

    SciTech Connect

    Myers, S C

    2001-02-05

    Calibration of seismic nuclear test monitoring stations relies on a diverse reference-event database. The reference-event covariance structure must be characterized and the covariance propagated to subsequent processes and calibration products. In seismic location one of the first steps in the calibration process is declustering, in which closely spaced epicenters are combined to reduce redundant data and random observational errors. We formulate a new declustering procedure that accounts for correlated and uncorrelated components of reference-event error, producing a declustered data set that tracks raw reference-event uncertainties. Declustering behavior is demonstrated using example data sets.

  19. Calibration of X-Ray diffractometer by the experimental comparison method

    SciTech Connect

    Dudka, A. P.

    2015-07-15

    A software for calibrating an X-ray diffractometer with area detector has been developed. It is proposed to search for detector and goniometer calibration models whose parameters are reproduced in a series of measurements on a reference crystal. Reference (standard) crystals are prepared during the investigation; they should provide the agreement of structural models in repeated analyses. The technique developed has been used to calibrate Xcalibur Sapphire and Eos, Gemini Ruby (Agilent) and Apex x8 and Apex Duo (Bruker) diffractometers. The main conclusions are as follows: the calibration maps are stable for several years and can be used to improve structural results, verified CCD detectors exhibit significant inhomogeneity of the efficiency (response) function, and a Bruker goniometer introduces smaller distortions than an Agilent goniometer.

  20. Calibration of X-Ray diffractometer by the experimental comparison method

    NASA Astrophysics Data System (ADS)

    Dudka, A. P.

    2015-07-01

    A software for calibrating an X-ray diffractometer with area detector has been developed. It is proposed to search for detector and goniometer calibration models whose parameters are reproduced in a series of measurements on a reference crystal. Reference (standard) crystals are prepared during the investigation; they should provide the agreement of structural models in repeated analyses. The technique developed has been used to calibrate Xcalibur Sapphire and Eos, Gemini Ruby (Agilent) and Apex x8 and Apex Duo (Bruker) diffractometers. The main conclusions are as follows: the calibration maps are stable for several years and can be used to improve structural results, verified CCD detectors exhibit significant inhomogeneity of the efficiency (response) function, and a Bruker goniometer introduces smaller distortions than an Agilent goniometer.

  1. Research on autofocusing method with automatic calibration for aerial camera based on imaging resolution

    NASA Astrophysics Data System (ADS)

    Zhao, Yu-liang; Zhao, Hong-qiang; Li, Shu-jun; Zhang, Yu-ye

    2014-09-01

    Air materiel depot is a warehouse which store consumed all the parts and equipment vault of the plane. In order to ensure the various aviation equipment integrity of the backup piece rate, the inside temperature of depot must be controlled within a certain range. Therefore, the depot must be equipped a self-contained temperature real-time monitoring system. This paper presents a distributed temperature sensing alarm system to apply to real-time measure spatial distribution of temperature field. In order to eliminate influence to the scattering strength from the light source instability and the fiber bending splice loss and to improve temperature measurement accuracy, the system design used dual-channel dual-wavelength comparison method which make Anti-Stokes as signal channel and Stokes as a reference channel to collect signals of two channel respectively and detect the ratio of the two channels' signals. The light of LD directional coupling to the sensing optical fiber in the temperature field to test, domain reflect light from the sensing optical fiber directional coupling to receive channel again, Rayleigh domain reflect light is filtered after optical filter, the Anti-Stokes and Stokes are both taken out, converted and magnified, the two signals is digitalized by A/D Converter, and written to the storage machine , which linear cumulative to the content of the storage unit, The distributed measurement of the temperature field to test is finished. The collected 2900 measuring points real-time on 2km of optical fiber. The spatial resolution of the system was 0.7m, measurement range was -20-370 °C, and measurement error was +/- 2 °C. All index of the system achieved the desired objective. To get an accurate temperature field spatial distribution and the information of temporal variation, the system enabled real-time temperature of aviation depot monitoring and early warning. As a new sensing technology, the distributed fiber optic sensor has the functions of self

  2. Light calibration and quality assessment methods for Reflectance Transformation Imaging applied to artworks' analysis

    NASA Astrophysics Data System (ADS)

    Giachetti, A.; Daffara, C.; Reghelin, C.; Gobbetti, E.; Pintus, R.

    2015-06-01

    In this paper we analyze some problems related to the acquisition of multiple illumination images for Polynomial Texture Maps (PTM) or generic Reflectance Transform Imaging (RTI). We show that intensity and directionality nonuniformity can be a relevant issue when acquiring manual sets of images with the standard highlight-based setup both using a flash lamp and a LED light. To maintain a cheap and flexible acquisition setup that can be used on field and by non-experienced users we propose to use a dynamic calibration and correction of the lights based on multiple intensity and direction estimation around the imaged object during the acquisition. Preliminary tests on the results obtained have been performed by acquiring a specifically designed 3D printed pattern in order to see the accuracy of the acquisition obtained both for spatial discrimination of small structures and normal estimation, and on samples of different types of paper in order to evaluate material discrimination. We plan to design and build from our analysis and from the tools developed and under development a set of novel procedures and guidelines that can be used to turn the cheap and common RTI acquisition setup from a simple way to enrich object visualization into a powerful method for extracting quantitative characterization both of surface geometry and of reflective properties of different materials. These results could have relevant applications in the Cultural Heritage domain, in order to recognize different materials used in paintings or investigate the ageing status of artifacts' surface.

  3. A novel method for the in situ calibration of flow effects on a phosphate passive sampler.

    PubMed

    Sara O'Brien, Dominique; Chiswell, Barry; Mueller, Jochen F

    2009-01-01

    Monitoring of nutrients including phosphate in the aquatic environment remains a challenge. In the last decade passive sampling techniques have been developed that facilitates the time integrated monitoring of phosphate (P) through the use of an iron hydroxide (ferrihydrite) to sequester dissolved phosphate from solution. These methods rely on established techniques to negate the effects of flow (and associated turbulence) and control the rate at which chemicals accumulate within passive samplers. In this study we present a phosphate sampler within which a suspension of ferrihydrite is contained behind a commercially available membrane. Accumulation of dissolved phosphates into the P-sampler is governed by the rate at which ions are diffusing through the membrane and the water boundary layer (WBL). As the WBL changes subject to flow we have adopted an in situ calibration technique based on the dissolution of gypsum to predict the change in the rate of uptake dependent on flow. Here we demonstrate that the loss of gypsum from the passive flow monitor (PFM) can be used to predict the sampling rate (the volume of water extracted per day) for phosphate as a function of water velocity. The outcome of this study presents a new in-field tool for more accurate prediction of the effect of flow/turbulence on the uptake kinetics into passive samplers that is controlled by the diffusion of the chemical of interest through the stagnant water boundary layer.

  4. Passive sampling methods for contaminated sediments: Practical guidance for selection, calibration, and implementation

    PubMed Central

    Ghosh, Upal; Driscoll, Susan Kane; Burgess, Robert M; Jonker, Michiel To; Reible, Danny; Gobas, Frank; Choi, Yongju; Apitz, Sabine E; Maruya, Keith A; Gala, William R; Mortimer, Munro; Beegan, Chris

    2014-01-01

    This article provides practical guidance on the use of passive sampling methods (PSMs) that target the freely dissolved concentration (Cfree) for improved exposure assessment of hydrophobic organic chemicals in sediments. Primary considerations for selecting a PSM for a specific application include clear delineation of measurement goals for Cfree, whether laboratory-based “ex situ” and/or field-based “in situ” application is desired, and ultimately which PSM is best-suited to fulfill the measurement objectives. Guidelines for proper calibration and validation of PSMs, including use of provisional values for polymer–water partition coefficients, determination of equilibrium status, and confirmation of nondepletive measurement conditions are defined. A hypothetical example is described to illustrate how the measurement of Cfree afforded by PSMs reduces uncertainty in assessing narcotic toxicity for sediments contaminated with polycyclic aromatic hydrocarbons. The article concludes with a discussion of future research that will improve the quality and robustness of Cfree measurements using PSMs, providing a sound scientific basis to support risk assessment and contaminated sediment management decisions. Integr Environ Assess Manag 2014;10:210–223. © 2014 The Authors. Integrated Environmental Assessment and Management published by Wiley Periodicals, Inc. on behalf of SETAC. PMID:24288273

  5. Technique for Radiometer and Antenna Array Calibration with a Radiated Noise Diode

    NASA Technical Reports Server (NTRS)

    Srinivasan, Karthik; Limaye, Ashutosh; Laymon, Charles; Meyer, Paul

    2009-01-01

    This paper presents a new technique to calibrate a microwave radiometer and antenna array system. This calibration technique uses a radiated noise source in addition to two calibration sources internal to the radiometer. The method accurately calibrates antenna arrays with embedded active devices (such as amplifiers) which are used extensively in active phased array antennas.

  6. Developing a Measure of Wealth for Primary Student Families in a Developing Country: Comparison of Two Methods of Psychometric Calibration

    ERIC Educational Resources Information Center

    Griffin, Patrick

    2005-01-01

    This article compares the invariance properties of two methods of psychometric instrument calibration for the development of a measure of wealth among families of Grade 5 pupils in five provinces in Vietnam. The measure is based on self-reported lists of possessions in the home. Its stability has been measured over two time periods. The concept of…

  7. The use of suitable pseudo-invariant targets for MIVIS data calibration by the empirical line method

    NASA Astrophysics Data System (ADS)

    Mei, Alessandro; Bassani, Cristiana; Fontinovo, Giuliano; Salvatori, Rosamaria; Allegrini, Alessia

    2016-04-01

    The Empirical Line Method (ELM) enables the calibration of multi- and hyper-airborne/satellite image converting DN or radiance to reflectance values performed by using at ground data. High quality outcome can be reached with the selection of appropriate Pseudo-Invariant Targets (PIT). In this paper, spectral variability of "usual" (asphalt and concrete) and "unusual" (calcareous gravel, basaltic paving, concrete bricks, tartan paving and artificial turf) PITs is retrieved for ELM application. Such PITs are used to calibrate the Multispectral Infrared and Visible Imaging Spectrometer (MIVIS) airborne sensor in 12 different Runs. Firstly, processing of field spectral data enables the evaluation of pseudo-invariance of targets by studying their spectral changes in space and in time. Finally, these surfaces are used as Ground Calibration (GCT) and Validation Targets (GVT) in ELM. High calibration accuracy values are observed in Visible (VIS) range (98.9%) while a general decrease of accuracy in Near-InfraRed (NIR) (96.6%) and Middle-InfraRed (SWIR) (88.1%) are reached. Outcomes show that "usual" surfaces as asphalt and concrete and "unusual" surfaces such as tartan can be successfully used for MIVIS image calibration.

  8. A Monte Carlo (MC) based individual calibration method for in vivo x-ray fluorescence analysis (XRF).

    PubMed

    Hansson, Marie; Isaksson, Mats

    2007-04-07

    X-ray fluorescence analysis (XRF) is a non-invasive method that can be used for in vivo determination of thyroid iodine content. System calibrations with phantoms resembling the neck may give misleading results in the cases when the measurement situation largely differs from the calibration situation. In such cases, Monte Carlo (MC) simulations offer a possibility of improving the calibration by better accounting for individual features of the measured subjects. This study investigates the prospects of implementing MC simulations in a calibration procedure applicable to in vivo XRF measurements. Simulations were performed with Penelope 2005 to examine a procedure where a parameter, independent of the iodine concentration, was used to get an estimate of the expected detector signal if the thyroid had been measured outside the neck. An attempt to increase the simulation speed and reduce the variance by exclusion of electrons and by implementation of interaction forcing was conducted. Special attention was given to the geometry features: analysed volume, source-sample-detector distances, thyroid lobe size and position in the neck. Implementation of interaction forcing and exclusion of electrons had no obvious adverse effect on the quotients while the simulation time involved in an individual calibration was low enough to be clinically feasible.

  9. The extended wedge method: Atomic force microscope friction calibration for improved tolerance to instrument misalignments, tip offset, and blunt probes

    SciTech Connect

    Khare, H. S.; Burris, D. L.

    2013-05-15

    One of the major challenges in understanding and controlling friction is the difficulty in bridging the length and time scales of macroscale contacts and those of the single asperity interactions they comprise. While the atomic force microscope (AFM) offers a unique ability to probe tribological surfaces in a wear-free single-asperity contact, instrument calibration challenges have limited the usefulness of this technique for quantitative nanotribological studies. A number of lateral force calibration techniques have been proposed and used, but none has gained universal acceptance due to practical considerations, configuration limitations, or sensitivities to unknowable error sources. This paper describes a simple extension of the classic wedge method of AFM lateral force calibration which: (1) allows simultaneous calibration and measurement on any substrate, thus eliminating prior tip damage and confounding effects of instrument setup adjustments; (2) is insensitive to adhesion, PSD cross-talk, transducer/piezo-tube axis misalignment, and shear-center offset; (3) is applicable to integrated tips and colloidal probes; and (4) is generally applicable to any reciprocating friction coefficient measurement. The method was applied to AFM measurements of polished carbon (99.999% graphite) and single crystal MoS{sub 2} to demonstrate the technique. Carbon and single crystal MoS{sub 2} had friction coefficients of {mu}= 0.20 {+-} 0.04 and {mu}= 0.006 {+-} 0.001, respectively, against an integrated Si probe. Against a glass colloidal sphere, MoS{sub 2} had a friction coefficient of {mu}= 0.005 {+-} 0.001. Generally, the measurement uncertainties ranged from 10%-20% and were driven by the effect of actual frictional variation on the calibration rather than calibration error itself (i.e., due to misalignment, tip-offset, or probe radius).

  10. Laser photolysis of caged compounds at 405 nm: photochemical advantages, localisation, phototoxicity and methods for calibration.

    PubMed

    Trigo, Federico F; Corrie, John E T; Ogden, David

    2009-05-30

    Rapid, localised photolytic release of neurotransmitters from caged precursors at synaptic regions in the extracellular space is greatly hampered at irradiation wavelengths in the near-UV, close to the wavelength of maximum absorption of the caged precursor, because of inner-filtering by strong absorption of light in the cage solution between the objective and cell. For this reason two-photon excitation is commonly used for photolysis, particularly at multiple points distributed over large fields; or, with near-UV, if combined with local perfusion of the cage. These methods each have problems: the small cross-sections of common cages with two-photon excitation require high cage concentrations and light intensities near the phototoxic limit, while local perfusion gives non-uniform cage concentrations over the field of view. Single-photon photolysis at 405 nm, although less efficient than at 330-350 nm, with present cages is more efficient than two-photon photolysis. The reduced light absorption in the bulk cage solution permits efficient wide-field uncaging at non-toxic intensities with uniform cage concentration. Full photolysis of MNI-glutamate with 100 micros pulses required intensities of 2 mW microm(-2) at the preparation, shown to be non-toxic with repeated exposures. Light scattering at 405 nm was estimated as 50% at 18 microm depth in 21-day rat cerebellum. Methods are described for: (1) varying the laser spot size; (2) photolysis calibration in the microscope with the caged fluorophore NPE-HPTS over the wavelength range 347-405 nm; and (3) determining the point-spread function of excitation. Furthermore, DM-Nitrophen photolysis at 405 nm was efficient for intracellular investigations of Ca2+-dependent processes.

  11. Absolute x-ray energy calibration over a wide energy range using a diffraction-based iterative method.

    PubMed

    Hong, Xinguo; Chen, Zhiqiang; Duffy, Thomas S

    2012-06-01

    In this paper, we report a method of precise and fast absolute x-ray energy calibration over a wide energy range using an iterative x-ray diffraction based method. Although accurate x-ray energy calibration is indispensable for x-ray energy-sensitive scattering and diffraction experiments, there is still a lack of effective methods to precisely calibrate energy over a wide range, especially when normal transmission monitoring is not an option and complicated micro-focusing optics are fixed in place. It is found that by using an iterative algorithm the x-ray energy is only tied to the relative offset of sample-to-detector distance, which can be readily varied with high precision of the order of 10(-5) -10(-6) spatial resolution using gauge blocks. Even starting with arbitrary initial values of 0.1 Å, 0.3 Å, and 0.4 Å, the iteration process converges to a value within 3.5 eV for 31.122 keV x-rays after three iterations. Different common diffraction standards CeO(2), Au, and Si show an energy deviation of 14 eV. As an application, the proposed method has been applied to determine the energy-sensitive first sharp diffraction peak of network forming GeO(2) glass at high pressure, exhibiting a distinct behavior in the pressure range of 2-4 GPa. Another application presented is pair distribution function measurement using calibrated high-energy x-rays at 82.273 keV. Unlike the traditional x-ray absorption-based calibration method, the proposed approach does not rely on any edges of specific elements, and is applicable to the hard x-ray region where no appropriate absorption edge is available.

  12. Standard Reference Materials (SRMs) for the Calibration and Validation of Analytical Methods for PCBs (as Aroclor Mixtures).

    PubMed

    Poster, Dianne L; Schantz, Michele M; Leigh, Stefan D; Wise, Stephen A

    2004-01-01

    Six Standard Reference Materials (SRMs®) have been prepared by the National Institute of Standards and Technology (NIST) for the determination of PCBs as different Aroclor mixtures in methanol. Six additional SRMs of the same Aroclors in transformer oil have also been prepared. Specifically, solutions of Aroclors 1016, 1232, 1242, 1254, and 1260 have been gravimetrically prepared (individually) in methanol and transformer oil, mixed, and transferred to amber glass ampoules in approximately 1.2 mL aliquots. Gas chromatography with electron capture detection (GC-ECD) has been used to verify the gravimetric data for each solution and transformer oil SRM. Liquid chromatography was used for the isolation of the Aroclors from the transformer oil SRMs prior to GC-ECD analysis. Separate calibration solutions and oils were prepared with Aroclor levels similar to those in each methanol solution and transformer oil SRM and were processed alongside the samples. The GC-ECD response of each Aroclor was monitored relative to internal standards that were added to the complex mixtures for quantification. The gravimetric concentrations of Aroclors 1242 and 1254 in methanol were also examined by the same method of analysis (GC-ECD) using several different sources of Aroclors and two different capillary GC columns: a 5 % phenyl methylpolysiloxane phase and a relatively non-polar phase. The preparation of the materials, the gas chromatographic results, and the certified concentration values for each Aroclor SRM are described in this paper.

  13. A likelihood method to cross-calibrate air-shower detectors

    NASA Astrophysics Data System (ADS)

    Dembinski, Hans Peter; Kégl, Balázs; Mariş, Ioana C.; Roth, Markus; Veberič, Darko

    2016-01-01

    We present a detailed statistical treatment of the energy calibration of hybrid air-shower detectors, which combine a surface detector array and a fluorescence detector, to obtain an unbiased estimate of the calibration curve. The special features of calibration data from air showers prevent unbiased results, if a standard least-squares fit is applied to the problem. We develop a general maximum-likelihood approach, based on the detailed statistical model, to solve the problem. Our approach was developed for the Pierre Auger Observatory, but the applied principles are general and can be transferred to other air-shower experiments, even to the cross-calibration of other observables. Since our general likelihood function is expensive to compute, we derive two approximations with significantly smaller computational cost. In the recent years both have been used to calibrate data of the Pierre Auger Observatory. We demonstrate that these approximations introduce negligible bias when they are applied to simulated toy experiments, which mimic realistic experimental conditions.

  14. Field calibration of blowfly-derived DNA against traditional methods for assessing mammal diversity in tropical forests.

    PubMed

    Lee, Ping-Shin; Gan, Han Ming; Clements, Gopalasamy Reuben; Wilson, John-James

    2016-11-01

    Mammal diversity assessments based on DNA derived from invertebrates have been suggested as alternatives to assessments based on traditional methods; however, no study has field-tested both approaches simultaneously. In Peninsular Malaysia, we calibrated the performance of mammal DNA derived from blowflies (Diptera: Calliphoridae) against traditional methods used to detect species. We first compared five methods (cage trapping, mist netting, hair trapping, scat collection, and blowfly-derived DNA) in a forest reserve with no recent reports of megafauna. Blowfly-derived DNA and mist netting detected the joint highest number of species (n = 6). Only one species was detected by multiple methods. Compared to the other methods, blowfly-derived DNA detected both volant and non-volant species. In another forest reserve, rich in megafauna, we calibrated blowfly-derived DNA against camera traps. Blowfly-derived DNA detected more species (n = 11) than camera traps (n = 9), with only one species detected by both methods. The rarefaction curve indicated that blowfly-derived DNA would continue to detect more species with greater sampling effort. With further calibration, blowfly-derived DNA may join the list of traditional field methods. Areas for further investigation include blowfly feeding and dispersal biology, primer biases, and the assembly of a comprehensive and taxonomically-consistent DNA barcode reference library.

  15. Joining direct and indirect inverse calibration methods to characterize karst, coastal aquifers

    NASA Astrophysics Data System (ADS)

    De Filippis, Giovanna; Foglia, Laura; Giudici, Mauro; Mehl, Steffen; Margiotta, Stefano; Negri, Sergio

    2016-04-01

    Parameter estimation is extremely relevant for accurate simulation of groundwater flow. Parameter values for models of large-scale catchments are usually derived from a limited set of field observations, which can rarely be obtained in a straightforward way from field tests or laboratory measurements on samples, due to a number of factors, including measurement errors and inadequate sampling density. Indeed, a wide gap exists between the local scale, at which most of the observations are taken, and the regional or basin scale, at which the planning and management decisions are usually made. For this reason, the use of geologic information and field data is generally made by zoning the parameter fields. However, pure zoning does not perform well in the case of fairly complex aquifers and this is particularly true for karst aquifers. In fact, the support of the hydraulic conductivity measured in the field is normally much smaller than the cell size of the numerical model, so it should be upscaled to a scale consistent with that of the numerical model discretization. Automatic inverse calibration is a valuable procedure to identify model parameter values by conditioning on observed, available data, limiting the subjective evaluations introduced with the trial-and-error technique. Many approaches have been proposed to solve the inverse problem. Generally speaking, inverse methods fall into two groups: direct and indirect methods. Direct methods allow determination of hydraulic conductivities from the groundwater flow equations which relate the conductivity and head fields. Indirect methods, instead, can handle any type of parameters, independently from the mathematical equations that govern the process, and condition parameter values and model construction on measurements of model output quantities, compared with the available observation data, through the minimization of an objective function. Both approaches have pros and cons, depending also on model complexity. For

  16. A Method to Calibrate the High-resolution Catania Astrophysical Observatory Spectropolarimeter

    NASA Astrophysics Data System (ADS)

    Leone, F.; Avila, G.; Bellassai, G.; Bruno, P.; Catalano, S.; Di Benedetto, R.; Di Stefano, A.; Gangi, M.; Giarrusso, M.; Greco, V.; Martinetti, E.; Miraglia, M.; Munari, M.; Pontoni, C.; Scalia, C.; Scuderi, S.; Spanó, P.

    2016-05-01

    The Catania Astrophysical Observatory Spectropolarimeter (CAOS) is a white-pupil cross-dispersed échelle spectrograph with a spectral resolution of up to R = 55,000 in the 375-1100 nm range in a single exposure, with complete coverage up to 856 nm. CAOS is linked to the 36-inch telescope, at Mount Etna Observatory, with a couple of 100 μm optical fibers and it achieves a signal-to-noise ratio better than 60 for a V = 10 mag star in one hour. CAOS is thermally stabilized in temperature within a 0.01 K rms, so that radial velocities are measured with a precision better than 100 m s-1 from a single spectral line. Linear and circular spectropolarimetric observations are possible by means of a Savart plate working in series with a half-wave and a quarter-wave retarder plate in the 376-850 nm range. As is usual for high-resolution spectropolarimeters, CAOS is suitable to measure all Stokes parameters across spectral lines and it cannot measure the absolute degree of polarization. Observations of unpolarized standard stars show that instrumental polarization is generally zero at 550 nm and can increase up to 3% at the other wavelengths. Since polarized and unpolarized standard stars are useless, we suggest a method to calibrate a high-resolution spectropolarimeter on the basis of the polarimetric properties of spectral lines formed in the presence of a magnetic field. As applied to CAOS, observations of magnetic chemically peculiar stars of the main sequence show that the cross-talk from linear to circular polarization is smaller than 0.4% and that conversion from circular to linear is less than 2.7%. Strength and wavelength dependences of cross-talk can be entirely ascribed, via numerical simulations, to the incorrect retardance of achromatic wave plates.

  17. 76 FR 5319 - Regulation of Fuel and Fuel Additives: Alternative Test Method for Olefins in Gasoline

    Federal Register 2010, 2011, 2012, 2013, 2014

    2011-01-31

    ... AGENCY 40 CFR Part 80 RIN 2060-AP17 Regulation of Fuel and Fuel Additives: Alternative Test Method for... correlated to the fuel parameter's respective EPA designated test method. These alternative test methods are... sections 114(a) and 301(a) of the CAA. Regulation of Fuel and Fuel Additives: Alternative Test Method...

  18. A non-contact, thermal noise based method for the calibration of lateral deflection sensitivity in atomic force microscopy

    SciTech Connect

    Mullin, Nic Hobbs, Jamie K.

    2014-11-15

    Calibration of lateral forces and displacements has been a long standing problem in lateral force microscopies. Recently, it was shown by Wagner et al. that the thermal noise spectrum of the first torsional mode may be used to calibrate the deflection sensitivity of the detector. This method is quick, non-destructive and may be performed in situ in air or liquid. Here we make a full quantitative comparison of the lateral inverse optical lever sensitivity obtained by the lateral thermal noise method and the shape independent method developed by Anderson et al. We find that the thermal method provides accurate results for a wide variety of rectangular cantilevers, provided that the geometry of the cantilever is suitable for torsional stiffness calibration by the torsional Sader method, in-plane bending of the cantilever may be eliminated or accounted for and that any scaling of the lateral deflection signal between the measurement of the lateral thermal noise and the measurement of the lateral deflection is eliminated or corrected for. We also demonstrate that the thermal method may be used to characterize the linearity of the detector signal as a function of position, and find a deviation of less than 8% for the instrument used.

  19. Linear model correction: A method for transferring a near-infrared multivariate calibration model without standard samples.

    PubMed

    Liu, Yan; Cai, Wensheng; Shao, Xueguang

    2016-12-05

    Calibration transfer is essential for practical applications of near infrared (NIR) spectroscopy because the measurements of the spectra may be performed on different instruments and the difference between the instruments must be corrected. For most of calibration transfer methods, standard samples are necessary to construct the transfer model using the spectra of the samples measured on two instruments, named as master and slave instrument, respectively. In this work, a method named as linear model correction (LMC) is proposed for calibration transfer without standard samples. The method is based on the fact that, for the samples with similar physical and chemical properties, the spectra measured on different instruments are linearly correlated. The fact makes the coefficients of the linear models constructed by the spectra measured on different instruments are similar in profile. Therefore, by using the constrained optimization method, the coefficients of the master model can be transferred into that of the slave model with a few spectra measured on slave instrument. Two NIR datasets of corn and plant leaf samples measured with different instruments are used to test the performance of the method. The results show that, for both the datasets, the spectra can be correctly predicted using the transferred partial least squares (PLS) models. Because standard samples are not necessary in the method, it may be more useful in practical uses.

  20. Linear model correction: A method for transferring a near-infrared multivariate calibration model without standard samples

    NASA Astrophysics Data System (ADS)

    Liu, Yan; Cai, Wensheng; Shao, Xueguang

    2016-12-01

    Calibration transfer is essential for practical applications of near infrared (NIR) spectroscopy because the measurements of the spectra may be performed on different instruments and the difference between the instruments must be corrected. For most of calibration transfer methods, standard samples are necessary to construct the transfer model using the spectra of the samples measured on two instruments, named as master and slave instrument, respectively. In this work, a method named as linear model correction (LMC) is proposed for calibration transfer without standard samples. The method is based on the fact that, for the samples with similar physical and chemical properties, the spectra measured on different instruments are linearly correlated. The fact makes the coefficients of the linear models constructed by the spectra measured on different instruments are similar in profile. Therefore, by using the constrained optimization method, the coefficients of the master model can be transferred into that of the slave model with a few spectra measured on slave instrument. Two NIR datasets of corn and plant leaf samples measured with different instruments are used to test the performance of the method. The results show that, for both the datasets, the spectra can be correctly predicted using the transferred partial least squares (PLS) models. Because standard samples are not necessary in the method, it may be more useful in practical uses.

  1. Rock magnetic properties of dusty olivine: comparison and calibration of non-heating paleointensity methods

    NASA Astrophysics Data System (ADS)

    Lappe, S. L.; Harrison, R. J.; Feinberg, J. M.

    2012-12-01

    The mechanism of chondrule formation is an important outstanding question in cosmochemistry. Magnetic signals recorded by Fe-Ni nanoparticles in chondrules could carry clues to their origin. Recently, research in this area has focused on 'dusty olivine' in ordinary chondrites as potential carriers of pre-accretionary remanence. Dusty olivine is characterised by the presence of sub-micron Fe-Ni inclusions within the olivine host. These metal particles form via subsolidus reduction of the olivine during chondrule formation and are thought to be protected from subsequent chemical and thermal alteration by the host olivine. Three sets of synthetic dusty olivines have been produced, using natural olivine (average Ni-content of 0.3 wt%), synthetic Ni-containing olivine (0.1wt% Ni) and synthetic Ni-free olivine as starting materials. The starting materials were ground to powders, packed into a 8-27 mm3 graphite crucible, heated up to 1350°C under a pure CO gas flow and kept at this temperature for 10 minutes. After this the samples were held in fixed orientation and quenched into water in a range of known magnetic fields from 0.2 mT to 1.5 mT. We present a comparison of all non-heating methods commonly used for paleointensity determination of extraterrestrial material. All samples showed uni-directional, single-component demagnetization behaviour. Saturation REM ratio (NRM/SIRM) and REMc ratio show non-linear behaviour as function of applied field and a saturation value < 1. Using the REM' method the samples showed approximately constant REM' between 100 and 150 mT AF-field. Plotting the average values for this field range again shows non-linear behaviour and a saturation value < 1. Another approach we examined to obtain calibration curves for paleointensity determination is based on ARM measurents. We also present an analysis of a new FORC-based method of paleointensity determination applied to metallic Fe-bearing samples [1, 2]. The method uses a first-order reversal

  2. Calibration of the pressure sensitivity of microphones by a free-field method at frequencies up to 80 khz.

    PubMed

    Zuckerwar, Allan J; Herring, G C; Elbing, Brian R

    2006-01-01

    A free-field (FF) substitution method for calibrating the pressure sensitivity of microphones at frequencies up to 80 kHz is demonstrated with both grazing and normal-incidence geometries. The substitution-based method, as opposed to a simultaneous method, avoids problems associated with the nonuniformity of the sound field and, as applied here, uses a 1/4-in. air-condenser pressure microphone as a known reference. Best results were obtained with a centrifugal fan, which is used as a random, broadband sound source. A broadband source minimizes reflection-related interferences that can plague FF measurements. Calibrations were performed on 1/4-in. FF air-condenser, electret, and microelectromechanical systems (MEMS) microphones in an anechoic chamber. The uncertainty of this FF method is estimated by comparing the pressure sensitivity of an air-condenser FF microphone, as derived from the FF measurement, with that of an electrostatic actuator calibration. The root-mean-square difference is found to be +/- 0.3 dB over the range 1-80 kHz, and the combined standard uncertainty of the FF method, including other significant contributions, is +/- 0.41 dB.

  3. A Standard Reference Material for Calibration of the Cup Furnace Smoke Toxicity Method for Assessing the Acute Inhalation Toxicity of Combustion Products.

    PubMed

    Levin, Barbara C; Paabo, Maya; Schiller, Susannah B

    1991-01-01

    A standard reference material (SRM 1048) has been developed for use with the cup furnace smoke toxicity method. This SRM is to be used to calibrate the apparatus and to enable the user to have confidence that the method is being conducted in a correct manner and that the equipment is functioning properly. The toxicological results from this SRM should not be used to compare with those of other materials (i.e., to determine if the combustion products of a test material are more or less toxic than those from this SRM). SRM 1048 is an acrylonitrile-butadiene-styrene (ABS) and is the same as SRM 1007B which is used for calibrating the flaming mode of the Smoke Density Chamber test method (ASTM E-662 and NFPA 258). For the purposes of calibrating the cup furnace smoke toxicity method, LC50 and N-Gas values plus their respective 95% confidence limits have been determined and certified for two combustion modes (flaming and nonflaming) and two observation periods (for the 30 min exposure only and for the 30 min exposure plus a 14 d post-exposure period). The certified LC50 values plus 95% confidence intervals (in g/m(3)) are 27 ± 3 (30 min, flaming); 25 ± 3 (30 min+ 14 d, flaming); 58 ± 15 (30 min, nonflaming); and 53 + 12 (30 min+ 14 d, nonflaming). The certified N-Gas values plus 95% confidence intervals are 1.4 ± 0.2 (30 min, flaming); 1.5 ± 0.2 (30 min+ 14 d, flaming); 1.2 ± 0.2 (30 min, nonflaming); and 1.4 ± 0.2 (30 min+ 14 d, nonflaming). It is recommended that this SRM be used with the N-Gas approach to calibrate the cup furnace smoke toxicity method rather than to determine the complete LC50 values. The N-Gas approach has the advantage of providing information on the gases responsible for the lethalities as well as the toxic potency of the smoke. In addition, the N-Gas approach reduces the number of experimental animals, the time necessary to complete the calibration, and the expense.

  4. Calibrating nonlinear volcano deformation source parameters in FEMs: The pinned mesh perturbation method. (Invited)

    NASA Astrophysics Data System (ADS)

    Masterlark, T.; Stone, J.; Feigl, K.

    2010-12-01

    The internal structure, loading processes, and effective boundary conditions of a volcano control the deformation that we observe at the Earth’s surface. Forward models of these internal structures and processes allow us to predict the surface deformation. In practice, we are faced with the inverse situation of using surface observations (e.g., InSAR and GPS) to characterize the inaccessible internal structures and processes. Distortions of these characteristics are tied to our ability to: 1) identify and resolve the internal structure; 2) simulate the internal processes over a problem domain having this internal structure; and 3) calibrate parameters that describe these internal processes to the observed deformation. Relatively simple analytical solutions for deformation sources (such as a pressurized magma chamber) embedded in a homogeneous, elastic half-space are commonly used to simulate observed volcano deformation, because they are computationally inexpensive, and thus easily integrated into inverse analyses that seek to characterize the source position and magnitude. However, the half-space models generally do not adequately represent internal distributions of material properties and complex geometric configurations, such as topography, of volcano deformational systems. These incompatibilities are known to severely bias both source parameter estimations and forward model calculations of deformation and stress. Alternatively, a Finite Element Model (FEM) can simulate the elastic response to a pressurized magma chamber over a domain having arbitrary geometry and distribution of material properties. However, the ability to impose perturbations of the source position parameters and automatically reconstruct an acceptable mesh has been an obstacle to implementing FEM-based nonlinear inverse methods to estimate the position of a deformation source. Using InSAR-observed deflation of Okmok volcano, Alaska, during its 1997 eruption as an example, we present the

  5. On the use of multi-algorithm, genetically adaptive multi-objective method for multi-site calibration of the SWAT model

    SciTech Connect

    Zhang, Xuesong; Srinivasan, Raghavan; Van Liew, M.

    2010-04-15

    With the availability of spatially distributed data, distributed hydrologic models are increasingly used for simulation of spatially varied hydrologic processes to understand and manage natural and human activities that affect watershed systems. Multi-objective optimization methods have been applied to calibrate distributed hydrologic models using observed data from multiple sites. As the time consumed by running these complex models is increasing substantially, selecting efficient and effective multi-objective optimization algorithms is becoming a nontrivial issue. In this study, we evaluated a multi-algorithm, genetically adaptive multi-objective method (AMALGAM) for multi-site calibration of a distributed hydrologic model—Soil and Water Assessment Tool (SWAT), and compared its performance with two widely used evolutionary multi-objective optimization (EMO) algorithms (i.e. Strength Pareto Evolutionary Algorithm 2 (SPEA2) and Non-dominated Sorted Genetic Algorithm II (NSGA-II)). In order to provide insights into each method’s overall performance, these three methods were tested in four watersheds with various characteristics. The test results indicate that the AMALGAM can consistently provide competitive or superior results compared with the other two methods. The multi-method search framework of AMALGAM, which can flexibly and adaptively utilize multiple optimization algorithms, makes it a promising tool for multi-site calibration of the distributed SWAT. For practical use of AMALGAM, it is suggested to implement this method in multiple trials with relatively small number of model runs rather than run it once with long iterations. In addition, incorporating different multiobjective optimization algorithms and multi-mode search operators into AMALGAM deserves further research.

  6. A novel plane method to the calibration of the thermal camera

    NASA Astrophysics Data System (ADS)

    Wang, Xunsi; Huang, Wei; Nie, Qiu-hua; Xu, Tiefeng; Dai, Shixun; Shen, Xiang; Cheng, Weihai

    2009-07-01

    This paper provides an up-to-date review of research efforts in thermal camera and target object recognition techniques based on two-dimensional (2D) images in the infrared (IR) spectra (8-12μm). From the geometric point of view, a special target plate was constructed with a radiation source of lamp excited that allows all of these devices to be calibrated geometrically along a radiance-based approach. The calibration theory and actual experimental procedures were described, then an automated measurement of the circle targets by image centroid algorithm. The key parameters of IR camera were calibrated out with 3 inner and 6 outer of Tsai model in thermal imaging. The subsequent data processing and analysis were then outlined. The 3D model from the successful calibration of a representative sample of the infrared array camera was presented and discussed. They provide much new and easy way to the geometric characteristics of these imagers that can be used in car-night-vision, medical, industrial, military, and environmental applications.

  7. A method for calibrating the relative gamma-ray light yield of plastic scintillators

    NASA Astrophysics Data System (ADS)

    Mengesha, W.; Feng, P. L.; Cordaro, J. G.; Anstey, M. R.; Myllenbeck, N. R.; Throckmorton, D. J.

    2017-03-01

    Currently we are investigating the inclusion of organotin compounds in new polystyrene scintillator materials to improve full gamma-ray energy sensitivity. Accurate calibration of the relative light yield from the newly developed scintillators is crucial to assess merits of compounds and chemical processes used in the scintillators' development. The full energy gamma-ray peak in a measured gamma-ray spectrum is commonly used in calibrating the relative light yield. However, the Compton continuum in the newly developed plastic scintillators is measured with much better efficiency and statistics and is found to be the best spectral feature that can be exploited for expeditious calibration of the relative light yield. In this study, we present a spectral gain matching of measured and simulated spectra, using a spectrum rebinning technique, to determine the Compton edge in a measured Compton continuum for accurate relative light yield calibration. The Compton edges determined using this technique were found to be within 1.2% of their theoretical estimates.

  8. Apparatus for measuring resistance change only in a cell analyzer and method for calibrating it

    DOEpatents

    Hoffman, Robert A.

    1980-01-01

    The disclosure relates to resistance only monitoring and calibration in an electrical cell analyzer. Sample and sheath fluid flows of different salinities are utilized, the sample flow being diameter modulated to produce a selected pattern which is compared to the resistance measured across the flows.

  9. Method and apparatus of a portable imaging-based measurement with self calibration

    DOEpatents

    Chang, Tzyy-Shuh [Ann Arbor, MI; Huang, Hsun-Hau [Ann Arbor, MI

    2012-07-31

    A portable imaging-based measurement device is developed to perform 2D projection based measurements on an object that is difficult or dangerous to access. This device is equipped with self calibration capability and built-in operating procedures to ensure proper imaging based measurement.

  10. Solving the robot-world, hand-eye(s) calibration problem with iterative methods

    Technology Transfer Automated Retrieval System (TEKTRAN)

    Robot-world, hand-eye calibration is the problem of determining the transformation between the robot end effector and a camera, as well as the transformation between the robot base and the world coordinate system. This relationship has been modeled as AX = ZB, where X and Z are unknown homogeneous ...

  11. Assessment of measurement errors and dynamic calibration methods for three different tipping bucket rain gauges

    Technology Transfer Automated Retrieval System (TEKTRAN)

    Three different models of tipping bucket rain gauges (TBRs), viz. HS-TB3 (Hydrological Services Pty Ltd), ISCO-674 (Isco, Inc.) and TR-525 (Texas Electronics, Inc.), were calibrated in the lab to quantify measurement errors across a range of rainfall intensities (5 mm.h-1 to 250 mm.h-1) and three di...

  12. Application of the Langley plot method to the calibration of the solar backscattered ultraviolet instrument on the Nimbus 7 satellite

    NASA Technical Reports Server (NTRS)

    Bhartia, P. K.; Taylor, S.; Mcpeters, R. D.; Wellemeyer, C.

    1995-01-01

    The concept of the well-known Langley plot technique, used for the calibration of ground-based instruments, has been generalized for application to satellite instruments. In polar regions, near summer solstice, the solar backscattered ultraviolet (SBUV) instrument on the Nimbus 7 satellite samples the same ozone field at widely different solar zenith angles. These measurements are compared to assess the long-term drift in the instrument calibration. Although the technique provides only a relative wavelength-to-wavelength calibration, it can be combined with existing techniques to determine the drift of the instrument at any wavelength. Using this technique, we have generated a 12-year data set of ozone vertical profiles from SBUV with an estimated accuracy of +/- 5% at 1 mbar and +/- 2% at 10 mbar (95% confidence) over 12 years. Since the method is insensitive to true changes in the atmospheric ozone profile, it can also be used to compare the calibrations of similar SBUV instruments launched without temporal overlap.

  13. TU-A-18A-01: Basic Principles of PET/CT, Calibration Methods and Contrast Recovery Across Multiple Cameras

    SciTech Connect

    Kappadath, S; Nye, J

    2014-06-15

    This continuing education session will discuss the physical principles of PET/CT imaging and characterization of contrast recovery using accreditation phantoms. A detailed overview will be given on the physical principles of PET including positron decay physics, 2D and 3D data acquisition, time-of-flight, scatter correction, CT attenuation correction, and image reconstruction. Instrument quality control and calibration procedures will be discussed. Technical challenges, common image artifacts and strategies to mitigate these issues will also be discussed. Data will be presented on acquisition techniques and reconstruction parameters affecting contrast recovery. The discussion will emphasize the minimization of reconstruction differences in quantification metrics such as SUV and contrast recovery coefficients for the NEMA and ACR clinical trial phantoms. Data from new and older generation scanners will be shown including comparison of contrast recovery measurements to their analytical solutions. The goal of this session is to update attendees on the quality control and calibration of PET/CT scanners, on methods to establish a common calibration for PET/CT scanners to control for instrument variance across multiple sites. Learning Objectives: Review the physical principles of PET/CT, quality control and calibration Gain further understanding on how to apply techniques for improving quantitative agreement across multiple cameras Describe the differences between measured and expected contrast recovery for the NEMA and ACR PET phantoms.

  14. The fractal calibration method applied to the characterization of polymers in solvent mixtures and in mixed gel packings by SEC.

    PubMed

    Porcar, Iolanda; García-Lopera, Rosa; Abad, Concepción; Campos, Agustín

    2007-08-01

    The size-exclusion chromatographic (SEC) behaviour of different solvent/polymer systems in three packing sets has been analysed from fractal considerations. The three-column sets studied are specifically formed by: (i) 'pure' micro-styragel, (ii) 'mixed' TSK Gel H(HR + XL + HR) and (iii) mixed TSK Gel H(XL + HR + XL). The experimental data reveals that in most of the systems assayed the classical universal calibration (UC) is not fulfilled, denoting the existence of secondary effects accompanying the main SEC mechanism. In order to obtain an accurate characterization of different polymers eluted in solvent mixtures and/or mixed packings, the use of a reliable and trusted calibration curve is required. In this sense, two alternative procedures have been analysed: the specific (SC) and the fractal (FC) calibrations. The results have evidenced that the use of the FC instead of the classical universal method diminishes up to nine times (in the case of the micro-styragel set) the mean deviation on the calculated molar mass with respect to the value given by the supplier. In the case of TSK Gel-based sets, the mean deviation is reduced to the half. The SC curve made with standards of the sample under study also reduces the mean deviation values but needs a broad set of narrow standards, whereas the fractal approach only needs one polymeric sample to build up the calibration curve.

  15. Additive Methods for Prediction of Thermochemical Properties. The Laidler Method Revisited. 1. Hydrocarbons

    NASA Astrophysics Data System (ADS)

    Leal, Joa˜O. Paulo

    2006-03-01

    A new parameterization of the Laidler method for estimation of atomization enthalpies and standard enthalpies of formation at 298.15 K for several families of hydrocarbons (alkanes, alkenes, alkynes, polyenes, poly-ynes, alkyl radicals, cycloalkanes, cycloalkenes, benzene derivatives, and polyaromatics) is presented. A total of 200 compounds (164 for liquid phase) are used for the calculation of the parameters. Comparison between the experimental values and those calculated using the group additive scheme led to an average difference of 1.28 kJṡmol-1 for the gas phase enthalpy of formation (excluding the polyaromatic compounds) and of 1.38 kJṡmol-1 for the liquid phase enthalpy of formation. The data base used appears to be essentially error free, but for some compounds (e.g., 2,2,4-trimethyl-pentane, with the highest deviation among all compounds except the polyaromatic ones) the experimental values might need a reevaluation. An Excel worksheet is provided to simplify the calculation of enthalpies of formation and atomization enthalpies based on the Laidler terms defined in this paper.

  16. Oil palm water use: calibration of a sap flux method and a field measurement scheme.

    PubMed

    Niu, Furong; Röll, Alexander; Hardanto, Afik; Meijide, Ana; Köhler, Michael; Hendrayanto; Hölscher, Dirk

    2015-05-01

    Oil palm (Elaeis guineensis Jacq.) water use was assessed by sap flux density measurements with the aim to establish the method and derive water-use characteristics. Thermal dissipation probes were inserted into leaf petioles of mature oil palms. In the laboratory, we tested our set-up against gravimetric measurements and derived new parameters for the original calibration equation that are specific to oil palm petioles. In the lowlands of Jambi, Indonesia, in a 12-year-old monoculture plantation, 56 leaves on 10 palms were equipped with one sensor per leaf. A 10-fold variation in individual leaf water use among leaves was observed, but we did not find significant correlations to the variables trunk height and diameter, leaf azimuthal orientation, leaf inclination or estimated horizontal leaf shading. We thus took an un-stratified approach to determine an appropriate sampling design to estimate stand transpiration (Es, mm day(-1)) rates of oil palm. We used the relative standard error of the mean (SEn, %) as a measure for the potential estimation error of Es associated with sample size. It was 14% for a sample size of 13 leaves to determine the average leaf water use and four palms to determine the average number of leaves per palm. Increasing these sample sizes only led to minor further decreases of the SEn of Es. The observed 90-day average of Es was 1.1 mm day(-1) (error margin ± 0.2 mm day(-1)), which seems relatively low, but does not contradict Penman-Monteith-derived estimates of evapotranspiration. Examining the environmental drivers of Es on an intra-daily scale indicates an early, pre-noon maximum of Es rates (11 am) due to a very sensitive reaction of Es to increasing vapor pressure deficit in the morning. This early peak is followed by a steady decline of Es rates for the rest of the day, despite further rising levels of vapor pressure deficit and radiation; this results in pronounced hysteresis, particularly between Es and vapor pressure deficit.

  17. Net analyte signal standard addition method for simultaneous determination of sulphadiazine and trimethoprim in bovine milk and veterinary medicines.

    PubMed

    Hajian, Reza; Mousavi, Esmat; Shams, Nafiseh

    2013-06-01

    Net analyte signal standard addition method has been used for the simultaneous determination of sulphadiazine and trimethoprim by spectrophotometry in some bovine milk and veterinary medicines. The method combines the advantages of standard addition method with the net analyte signal concept which enables the extraction of information concerning a certain analyte from spectra of multi-component mixtures. This method has some advantages such as the use of a full spectrum realisation, therefore it does not require calibration and prediction step and only a few measurements require for the determination. Cloud point extraction based on the phenomenon of solubilisation used for extraction of sulphadiazine and trimethoprim in bovine milk. It is based on the induction of micellar organised media by using Triton X-100 as an extraction solvent. At the optimum conditions, the norm of NAS vectors increased linearly with concentrations in the range of 1.0-150.0 μmolL(-1) for both sulphadiazine and trimethoprim. The limits of detection (LOD) for sulphadiazine and trimethoprim were 0.86 and 0.92 μmolL(-1), respectively.

  18. Improving self-calibration.

    PubMed

    Enßlin, Torsten A; Junklewitz, Henrik; Winderling, Lars; Greiner, Maksim; Selig, Marco

    2014-10-01

    Response calibration is the process of inferring how much the measured data depend on the signal one is interested in. It is essential for any quantitative signal estimation on the basis of the data. Here, we investigate self-calibration methods for linear signal measurements and linear dependence of the response on the calibration parameters. The common practice is to augment an external calibration solution using a known reference signal with an internal calibration on the unknown measurement signal itself. Contemporary self-calibration schemes try to find a self-consistent solution for signal and calibration by exploiting redundancies in the measurements. This can be understood in terms of maximizing the joint probability of signal and calibration. However, the full uncertainty structure of this joint probability around its maximum is thereby not taken into account by these schemes. Therefore, better schemes, in sense of minimal square error, can be designed by accounting for asymmetries in the uncertainty of signal and calibration. We argue that at least a systematic correction of the common self-calibration scheme should be applied in many measurement situations in order to properly treat uncertainties of the signal on which one calibrates. Otherwise, the calibration solutions suffer from a systematic bias, which consequently distorts the signal reconstruction. Furthermore, we argue that nonparametric, signal-to-noise filtered calibration should provide more accurate reconstructions than the common bin averages and provide a new, improved self-calibration scheme. We illustrate our findings with a simplistic numerical example.

  19. Simplified stereo-optical ultrasound plane calibration

    NASA Astrophysics Data System (ADS)

    Hoßbach, Martin; Noll, Matthias; Wesarg, Stefan

    2013-03-01

    Image guided therapy is a natural concept and commonly used in medicine. In anesthesia, a common task is the injection of an anesthetic close to a nerve under freehand ultrasound guidance. Several guidance systems exist using electromagnetic tracking of the ultrasound probe as well as the needle, providing the physician with a precise projection of the needle into the ultrasound image. This, however, requires additional expensive devices. We suggest using optical tracking with miniature cameras attached to a 2D ultrasound probe to achieve a higher acceptance among physicians. The purpose of this paper is to present an intuitive method to calibrate freehand ultrasound needle guidance systems employing a rigid stereo camera system. State of the art methods are based on a complex series of error prone coordinate system transformations which makes them susceptible to error accumulation. By reducing the amount of calibration steps to a single calibration procedure we provide a calibration method that is equivalent, yet not prone to error accumulation. It requires a linear calibration object and is validated on three datasets utilizing di erent calibration objects: a 6mm metal bar and a 1:25mm biopsy needle were used for experiments. Compared to existing calibration methods for freehand ultrasound needle guidance systems, we are able to achieve higher accuracy results while additionally reducing the overall calibration complexity. Ke

  20. Measurements of Temperature Variations in the Atmosphere near the Tropopause with Reference to Airspeed Calibration by the Temperature Method

    NASA Technical Reports Server (NTRS)

    Lina, Lindsay J; Ricker, Harry H , Jr

    1952-01-01

    Measurements of temperature variations in the atmosphere near the tropopause over land in the vicinity of Langley Field, Va., are presented. This investigation was made for the purpose of obtaining information on the accuracy of the temperature method(NACA TN 2046) of airspeed calibration over the range of Mach number from 0.6 to .08. The temperature surveys and the description of a specially designed thermometer are also presented.

  1. Biological chromodynamics: a general method for measuring protein occupancy across the genome by calibrating ChIP-seq

    PubMed Central

    Hu, Bin; Petela, Naomi; Kurze, Alexander; Chan, Kok-Lung; Chapard, Christophe; Nasmyth, Kim

    2015-01-01

    Sequencing DNA fragments associated with proteins following in vivo cross-linking with formaldehyde (known as ChIP-seq) has been used extensively to describe the distribution of proteins across genomes. It is not widely appreciated that this method merely estimates a protein's distribution and cannot reveal changes in occupancy between samples. To do this, we tagged with the same epitope orthologous proteins in Saccharomyces cerevisiae and Candida glabrata, whose sequences have diverged to a degree that most DNA fragments longer than 50 bp are unique to just one species. By mixing defined numbers of C. glabrata cells (the calibration genome) with S. cerevisiae samples (the experimental genomes) prior to chromatin fragmentation and immunoprecipitation, it is possible to derive a quantitative measure of occupancy (the occupancy ratio – OR) that enables a comparison of occupancies not only within but also between genomes. We demonstrate for the first time that this ‘internal standard’ calibration method satisfies the sine qua non for quantifying ChIP-seq profiles, namely linearity over a wide range. Crucially, by employing functional tagged proteins, our calibration process describes a method that distinguishes genuine association within ChIP-seq profiles from background noise. Our method is applicable to any protein, not merely highly conserved ones, and obviates the need for the time consuming, expensive, and technically demanding quantification of ChIP using qPCR, which can only be performed on individual loci. As we demonstrate for the first time in this paper, calibrated ChIP-seq represents a major step towards documenting the quantitative distributions of proteins along chromosomes in different cell states, which we term biological chromodynamics. PMID:26130708

  2. A New Calibration Method Using Low Cost MEM IMUs to Verify the Performance of UAV-Borne MMS Payloads

    PubMed Central

    Chiang, Kai-Wei; Tsai, Meng-Lun; Naser, El-Sheimy; Habib, Ayman; Chu, Chien-Hsun

    2015-01-01

    Spatial information plays a critical role in remote sensing and mapping applications such as environment surveying and disaster monitoring. An Unmanned Aerial Vehicle (UAV)-borne mobile mapping system (MMS) can accomplish rapid spatial information acquisition under limited sky conditions with better mobility and flexibility than other means. This study proposes a long endurance Direct Geo-referencing (DG)-based fixed-wing UAV photogrammetric platform and two DG modules that each use different commercial Micro-Electro Mechanical Systems’ (MEMS) tactical grade Inertial Measurement Units (IMUs). Furthermore, this study develops a novel kinematic calibration method which includes lever arms, boresight angles and camera shutter delay to improve positioning accuracy. The new calibration method is then compared with the traditional calibration approach. The results show that the accuracy of the DG can be significantly improved by flying at a lower altitude using the new higher specification hardware. The new proposed method improves the accuracy of DG by about 20%. The preliminary results show that two-dimensional (2D) horizontal DG positioning accuracy is around 5.8 m at a flight height of 300 m using the newly designed tactical grade integrated Positioning and Orientation System (POS). The positioning accuracy in three-dimensions (3D) is less than 8 m. PMID:25808764

  3. New calibration method using low cost MEM IMUs to verify the performance of UAV-borne MMS payloads.

    PubMed

    Chiang, Kai-Wei; Tsai, Meng-Lun; Naser, El-Sheimy; Habib, Ayman; Chu, Chien-Hsun

    2015-03-19

    Spatial information plays a critical role in remote sensing and mapping applications such as environment surveying and disaster monitoring. An Unmanned Aerial Vehicle (UAV)-borne mobile mapping system (MMS) can accomplish rapid spatial information acquisition under limited sky conditions with better mobility and flexibility than other means. This study proposes a long endurance Direct Geo-referencing (DG)-based fixed-wing UAV photogrammetric platform and two DG modules that each use different commercial Micro-Electro Mechanical Systems' (MEMS) tactical grade Inertial Measurement Units (IMUs). Furthermore, this study develops a novel kinematic calibration method which includes lever arms, boresight angles and camera shutter delay to improve positioning accuracy. The new calibration method is then compared with the traditional calibration approach. The results show that the accuracy of the DG can be significantly improved by flying at a lower altitude using the new higher specification hardware. The new proposed method improves the accuracy of DG by about 20%. The preliminary results show that two-dimensional (2D) horizontal DG positioning accuracy is around 5.8 m at a flight height of 300 m using the newly designed tactical grade integrated Positioning and Orientation System (POS). The positioning accuracy in three-dimensions (3D) is less than 8 m.

  4. White-light Interferometry using a Channeled Spectrum: II. Calibration Methods, Numerical and Experimental Results

    NASA Technical Reports Server (NTRS)

    Zhai, Chengxing; Milman, Mark H.; Regehr, Martin W.; Best, Paul K.

    2007-01-01

    In the companion paper, [Appl. Opt. 46, 5853 (2007)] a highly accurate white light interference model was developed from just a few key parameters characterized in terms of various moments of the source and instrument transmission function. We develop and implement the end-to-end process of calibrating these moment parameters together with the differential dispersion of the instrument and applying them to the algorithms developed in the companion paper. The calibration procedure developed herein is based on first obtaining the standard monochromatic parameters at the pixel level: wavenumber, phase, intensity, and visibility parameters via a nonlinear least-squares procedure that exploits the structure of the model. The pixel level parameters are then combined to obtain the required 'global' moment and dispersion parameters. The process is applied to both simulated scenarios of astrometric observations and to data from the microarcsecond metrology testbed (MAM), an interferometer testbed that has played a prominent role in the development of this technology.

  5. Dental haptic simulator to train hand skill of student - calibration method to realize visual haptic environment.

    PubMed

    Yamaguchi, Satoshi; Kazumichi, Wakabayashi; Yoshida, Yoshinori; Nagashima, Tadashi; Takeshige, Fumio; Kawamoto, Yusuke; Noborio, Hiroshi; Sohmura, Taiji

    2009-01-01

    In prior work, we have developed dental training simulator to train hand skill of student (HHDTS). In the present study, we performed calibration between haptic device coordinate system and half-mirror coordinate system for our system to realize real clinical situation. As a result, the user can overlay the CG (turbine) onto the stylus of haptic device and intract to the CG (tooth) directly.

  6. Probabilistic methods for sensitivity analysis and calibration in the NASA challenge problem

    DOE PAGES

    Safta, Cosmin; Sargsyan, Khachik; Najm, Habib N.; ...

    2015-01-01

    In this study, a series of algorithms are proposed to address the problems in the NASA Langley Research Center Multidisciplinary Uncertainty Quantification Challenge. A Bayesian approach is employed to characterize and calibrate the epistemic parameters based on the available data, whereas a variance-based global sensitivity analysis is used to rank the epistemic and aleatory model parameters. A nested sampling of the aleatory–epistemic space is proposed to propagate uncertainties from model parameters to output quantities of interest.

  7. Gamma spectrometry efficiency calibration using Monte Carlo methods to measure radioactivity of 137Cs in food samples.

    PubMed

    Alrefae, T

    2014-12-01

    A simple method of efficiency calibration for gamma spectrometry was performed. This method, which focused on measuring the radioactivity of (137)Cs in food samples, was based on Monte Carlo simulations available in the free-of-charge toolkit GEANT4. Experimentally, the efficiency values of a high-purity germanium detector were calculated for three reference materials representing three different food items. These efficiency values were compared with their counterparts produced by a computer code that simulated experimental conditions. Interestingly, the output of the simulation code was in acceptable agreement with the experimental findings, thus validating the proposed method.

  8. [Comparison of four multivariate calibration methods in simultaneous determination of air toxic organic compounds with FTIR spectroscopy].

    PubMed

    Li, Yan; Wang, Jun-de; Chen, Zuo-ru; Zhou, Xue-tie; Huang, Zhong-hua

    2002-10-01

    The concentration determination abilities of four multivariate calibration methods--classical least squares (CLS), partial least squares (PLS), kalman filter method (KFM) and artificial neural network (ANN) were compared in this paper. Five air toxic organic compounds--1,3-butadiene, benzene, o-xylen, chlorobenzene, and acrolein--whose FTIR spectra seriously overlap each other were selected to compose the analytical objects. The evaluation criterion was according to the mean prediction error (MPE) and mean relative error (MRE). Results showed that PLS was superior to other methods when treating multicomponent analysis problem, while there was no comparable difference between CLS, KFM and ANN.

  9. In situ broadband cryogenic calibration for two-port superconducting microwave resonators.

    PubMed

    Yeh, Jen-Hao; Anlage, Steven M

    2013-03-01

    We introduce an improved microwave calibration method for use in a cryogenic environment, based on a traditional three-standard calibration, the Thru-Reflect-Line (TRL) calibration. The modified calibration method takes advantage of additional information from multiple measurements of an ensemble of realizations of a superconducting resonator, as a new pseudo-Open standard, to correct errors in the TRL calibration. We also demonstrate an experimental realization of this in situ broadband cryogenic calibration system utilizing cryogenic switches. All calibration measurements are done in the same thermal cycle as the measurement of the resonator (requiring only an additional 20 min), thus avoiding 4 additional thermal cycles for traditional TRL calibration (which would require an additional 12 days). The experimental measurements on a wave-chaotic microwave billiard verify that the new method significantly improves the measured scattering matrix of a high-quality-factor superconducting resonator.

  10. Non-contact thrust stand calibration method for repetitively pulsed electric thrusters.

    PubMed

    Wong, Andrea R; Toftul, Alexandra; Polzin, Kurt A; Pearson, J Boise

    2012-02-01

    A thrust stand calibration technique for use in testing repetitively pulsed electric thrusters for in-space propulsion has been developed and tested using a modified hanging pendulum thrust stand. In the implementation of this technique, current pulses are applied to a solenoid to produce a pulsed magnetic field that acts against a permanent magnet mounted to the thrust stand pendulum arm. The force on the magnet is applied in this non-contact manner, with the entire pulsed force transferred to the pendulum arm through a piezoelectric force transducer to provide a time-accurate force measurement. Modeling of the pendulum arm dynamics reveals that after an initial transient in thrust stand motion the quasi-steady average deflection of the thrust stand arm away from the unforced or "zero" position can be related to the average applied force through a simple linear Hooke's law relationship. Modeling demonstrates that this technique is universally applicable except when the pulsing period is increased to the point where it approaches the period of natural thrust stand motion. Calibration data were obtained using a modified hanging pendulum thrust stand previously used for steady-state thrust measurements. Data were obtained for varying impulse bit at constant pulse frequency and for varying pulse frequency. The two data sets exhibit excellent quantitative agreement with each other. The overall error on the linear regression fit used to determine the calibration coefficient was roughly 1%.

  11. Method and apparatus for self-calibration and phasing of array antenna

    NASA Technical Reports Server (NTRS)

    Wu, C. (Inventor)

    1984-01-01

    A technique for self-calibrating and phasing a lens-feed array antenna, while normal operation is stopped, utilizes reflected energy of a continuous and coherent wave broadcast by a transmitter through a central feed while a phase controller advances the phase angles of reciprocal phase shifters in radiation electronics of the array elements at different rates to provide a distinct frequency modulation of electromagnetic wave energy returned by reflection in one mode and leakage in another mode from the radiation electronics of each array element. The composite return signal received by a synchronous receiver goes through a Fourier transform processing system and produces a response function for each antenna element. Compensation of the phase angles for the antenna elements required to conform the antenna response to a precomputed array pattern is derived from the reciprocal square root of the response functions for the antenna elements which, for a rectangular array of NXM elements, is a response function T(n,m). A third mode of calibration uses an external pilot tone from a separate antenna element. Respective responses are thus obtained from the three modes of calibration.

  12. Using structural equation modeling to construct calibration equations relating PM2.5 mass concentration samplers to the federal reference method sampler

    NASA Astrophysics Data System (ADS)

    Bilonick, Richard A.; Connell, Daniel P.; Talbott, Evelyn O.; Rager, Judith R.; Xue, Tao

    2015-02-01

    The objective of this study was to remove systematic bias among fine particulate matter (PM2.5) mass concentration measurements made by different types of samplers used in the Pittsburgh Aerosol Research and Inhalation Epidemiology Study (PARIES). PARIES is a retrospective epidemiology study that aims to provide a comprehensive analysis of the associations between air quality and human health effects in the Pittsburgh, Pennsylvania, region from 1999 to 2008. Calibration was needed in order to minimize the amount of systematic error in PM2.5 exposure estimation as a result of including data from 97 different PM2.5 samplers at 47 monitoring sites. Ordinary regression often has been used for calibrating air quality measurements from pairs of measurement devices; however, this is only appropriate when one of the two devices (the "independent" variable) is free from random error, which is rarely the case. A group of methods known as "errors-in-variables" (e.g., Deming regression, reduced major axis regression) has been developed to handle calibration between two devices when both are subject to random error, but these methods require information on the relative sizes of the random errors for each device, which typically cannot be obtained from the observed data. When data from more than two devices (or repeats of the same device) are available, the additional information is not used to inform the calibration. A more general approach that often has been overlooked is the use of a measurement error structural equation model (SEM) that allows the simultaneous comparison of three or more devices (or repeats). The theoretical underpinnings of all of these approaches to calibration are described, and the pros and cons of each are discussed. In particular, it is shown that both ordinary regression (when used for calibration) and Deming regression are particular examples of SEMs but with substantial deficiencies. To illustrate the use of SEMs, the 7865 daily average PM2.5 mass

  13. SU-F-BRA-09: New Efficient Method for Xoft Axxent Electronic Brachytherapy Source Calibration by Pre-Characterizing Surface Applicators

    SciTech Connect

    Pai, S

    2015-06-15

    Purpose: The objective is to improve the efficiency and efficacy of Xoft™ Axxent™ electronic brachytherapy (EBT) calibration of the source & surface applicator using AAPM TG-61 formalism. Methods: Current method of Xoft EBT source calibration involves determination of absolute dose rate of the source in each of the four conical surface applicators using in-air chamber measurements & TG61 formalism. We propose a simplified TG-61 calibration methodology involving initial characterization of surface cone applicators. This is accomplished by calibrating dose rates for all 4 surface applicator sets (for 10 sources) which establishes the “applicator output ratios” with respect to the selected reference applicator (20 mm applicator). After the initial time, Xoft™ Axxent™ source TG61 Calibration is carried out only in the reference applicator. Using the established applicator output ratios, dose rates for other applicators will be calculated. Results: 200 sources & 8 surface applicator sets were calibrated cumulatively using a Standard Imaging A20 ion-chamber in accordance with manufacturer-recommended protocols. Dose rates of 10, 20, 35 & 50mm applicators were normalized to the reference (20mm) applicator. The data in Figure 1 indicates that the normalized dose rate variation for each applicator for all 200 sources is better than ±3%. The average output ratios are 1.11, 1.02 and 0.49 for the 10 mm,35 mm and 50 mm applicators, respectively, which are in good agreement with the manufacturer’s published output ratios of 1.13, 1.02 and 0.49. Conclusion: Our measurements successfully demonstrate the accuracy of a new calibration method using a single surface applicator for Xoft EBT sources and deriving the dose rates of other applicators. The accuracy of the calibration is improved as this method minimizes the source position variation inside the applicator during individual source calibrations. The new method significantly reduces the calibration time to less

  14. Hyper-Cam automated calibration method for continuous hyperspectral imaging measurements

    NASA Astrophysics Data System (ADS)

    Gagnon, Jean-Philippe; Habte, Zewdu; George, Jacks; Farley, Vincent; Tremblay, Pierre; Chamberland, Martin; Romano, Joao; Rosario, Dalton

    2010-04-01

    The midwave and longwave infrared regions of the electromagnetic spectrum contain rich information which can be captured by hyperspectral sensors thus enabling enhanced detection of targets of interest. A continuous hyperspectral imaging measurement capability operated 24/7 over varying seasons and weather conditions permits the evaluation of hyperspectral imaging for detection of different types of targets in real world environments. Such a measurement site was built at Picatinny Arsenal under the Spectral and Polarimetric Imagery Collection Experiment (SPICE), where two Hyper-Cam hyperspectral imagers are installed at the Precision Armament Laboratory (PAL) and are operated autonomously since Fall of 2009. The Hyper-Cam are currently collecting a complete hyperspectral database that contains the MWIR and LWIR hyperspectral measurements of several targets under day, night, sunny, cloudy, foggy, rainy and snowy conditions. The Telops Hyper-Cam sensor is an imaging spectrometer that enables the spatial and spectral analysis capabilities using a single sensor. It is based on the Fourier-transform technology yielding high spectral resolution and enabling high accuracy radiometric calibration. It provides datacubes of up to 320x256 pixels at spectral resolutions of up to 0.25 cm-1. The MWIR version covers the 3 to 5 μm spectral range and the LWIR version covers the 8 to 12 μm spectral range. This paper describes the automated operation of the two Hyper-Cam sensors being used in the SPICE data collection. The Reveal Automation Control Software (RACS) developed collaboratively between Telops, ARDEC, and ARL enables flexible operating parameters and autonomous calibration. Under the RACS software, the Hyper-Cam sensors can autonomously calibrate itself using their internal blackbody targets, and the calibration events are initiated by user defined time intervals and on internal beamsplitter temperature monitoring. The RACS software is the first software developed for

  15. Analysis of Lard in Lipstick Formulation Using FTIR Spectroscopy and Multivariate Calibration: A Comparison of Three Extraction Methods.

    PubMed

    Waskitho, Dri; Lukitaningsih, Endang; Sudjadi; Rohman, Abdul

    2016-01-01

    Analysis of lard extracted from lipstick formulation containing castor oil has been performed using FTIR spectroscopic method combined with multivariate calibration. Three different extraction methods were compared, namely saponification method followed by liquid/liquid extraction with hexane/dichlorometane/ethanol/water, saponification method followed by liquid/liquid extraction with dichloromethane/ethanol/water, and Bligh & Dyer method using chloroform/methanol/water as extracting solvent. Qualitative and quantitative analysis of lard were performed using principle component (PCA) and partial least square (PLS) analysis, respectively. The results showed that, in all samples prepared by the three extraction methods, PCA was capable of identifying lard at wavelength region of 1200-800 cm(-1) with the best result was obtained by Bligh & Dyer method. Furthermore, PLS analysis at the same wavelength region used for qualification showed that Bligh and Dyer was the most suitable extraction method with the highest determination coefficient (R(2)) and the lowest root mean square error of calibration (RMSEC) as well as root mean square error of prediction (RMSEP) values.

  16. Calibration of a dual-PTZ-camera system for stereo vision based on parallel particle swarm optimization method

    NASA Astrophysics Data System (ADS)

    Chang, Yau-Zen; Wang, Huai-Ming; Lee, Shih-Tseng; Wu, Chieh-Tsai; Hsu, Ming-Hsi

    2014-02-01

    This work investigates the calibration of a stereo vision system based on two PTZ (Pan-Tilt-Zoom) cameras. As the accuracy of the system depends not only on intrinsic parameters, but also on the geometric relationships between rotation axes of the cameras, the major concern is the development of an effective and systematic way to obtain these relationships. We derived a complete geometric model of the dual-PTZ-camera system and proposed a calibration procedure for the intrinsic and external parameters of the model. The calibration method is based on Zhang's approach using an augmented checkerboard composed of eight small checkerboards, and is formulated as an optimization problem to be solved by an improved particle swarm optimization (PSO) method. Two Sony EVI-D70 PTZ cameras were used for the experiments. The root-mean-square errors (RMSE) of corner distances in the horizontal and vertical direction are 0.192 mm and 0.115 mm, respectively. The RMSE of overlapped points between the small checkerboards is 1.3958 mm.

  17. Improved 3D displacement measurements method and calibration of a combined fringe projection and 2D-DIC system

    NASA Astrophysics Data System (ADS)

    Siegmann, Philip; Felipe-Sese, Luis; Diaz-Garrido, Francisco

    2017-01-01

    An improved measurement method and an automatic calibration procedure are proposed for a combined 2D Digital Image Correlation and Fringe Projection system that allows measuring in- and out-of-plane displacement maps with only one image at each deformation stage of a specimen. The proposed method increases the accuracy and range of the out-of-plane displacements by taking into account the divergences of both the projected fringes (uncollimated) and the camera (with non-zero FOV). The calibration is performed automatically by acquiring a sequence of images of a reference plane by displacing perpendicular to it the camera and fringe projector with a motorized translation stage. The acquired images are then used to obtain a fringe function for each pixel and the necessary parameters required for the correction of the in-plane displacements. Furthermore, a closed form expression is obtained that relates the out-of-plane displacements with the shifted phase at each pixel for a given experimental set-up. This expression is in good agreement with the fringe function obtained by fitting a simple 2nd order polynomial to the experimental obtained calibration data. Finally, the polynomial approach is proposed as a fringe function because it avoids the errors in the determination of the required parameters of the theoretical expression as well as some small misalignment or aberration effects.

  18. Improving focus performance at litho using diffraction-based focus metrology, novel calibration methods, interface, and control loop

    NASA Astrophysics Data System (ADS)

    Hu, Jiarui; Chen, Y. L.; Chen, K. H.; Lee, Brian; Tsai, Frankie; Ke, C. M.; Liao, C. H.; Ngo, Desmond; Gosali, Benny; Tijssen, Robin; Huang, Vincent; Tu, Ward; Noot, Marc; Escalante Marun, Maryana; Leewis, Christian; Luijten, Carlo; Staals, Frank; Van Veen, Martijn; Furthner, Francois; Young, Stuart; Bhattacharyya, Kaustuve

    2016-03-01

    In advanced optical lithography the requirements of focus control continues to tighten. Usable depth of focus (DoF) is already quite low due to typical sources of focus errors, such as topography, wafer warpage and the thickness of photoresist. And now the usable DoF is further decreased by hotspots (design and imaging hotspots). All these have put extra challenges to improve focus metrology, scanner focus stability calibrations and on-product correction mechanisms. Asymmetric focus targets are developed to address robustness in focus measurements using diffraction-based focus (DBF and μDBF) metrology. A new layout specific calibration methodology is introduced for baseline focus setup and control in order to improve scanner focus uniformity and stability using the measurements of the above mentioned asymmetric targets. A similar metrology is also used for on product focus measurements. Moreover, a few novel alternative methods are also investigated for on-product focus measurements. Data shows good correlation between DBF and process on record (POR) method using traditional FEM. The new focus calibration demonstrated robustness, stability and speed. This technical publication will report the data from all the above activities including results from various product layers.

  19. [Construction and simplification of the calibration model for spectral analysis of fuel oil properties based on mutual information method].

    PubMed

    Hao, Yong; Sun, Xu-dong; Cai, Li-jun; Liu, Yan-de

    2012-01-01

    Near infrared diffuse reflectance (NIRS) and ultraviolet (UV) spectral analysis were adopted for quantitative determination of octane number and monoaromatics in fuel oil. Partial least squares regression (PLSR) was used for construction of vibrational spectral calibration models. Variables selection strategy based on mutual information (MI) theory was introduced to optimize the models for improving the precision and reducing the complexity. The results indicate that MI-PLSR method can effectively improve the predictive ability of the models and simplify them. For octane number models, the root mean square error of prediction (RMSEP) and the number of calibration variables were reduced from 0.288 and 401 to 0.111 and 112, respectively, and correlation coefficient (R) was improved from 0.985 to 0.998. For monoaromatics models, RMSEP and the number of calibration variables were reduced from 0.753 and 572 to 0.478 and 37, respectively, and R was improved from 0.996 to 0.998. Vibrational spectral analysis combined with MI-PLSR method can be used for quantitative analysis of fuel oil properties, and improve the cost-effectiveness.

  20. 40 CFR 80.8 - Sampling methods for gasoline, diesel fuel, fuel additives, and renewable fuels.

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... fuel, fuel additives, and renewable fuels. 80.8 Section 80.8 Protection of Environment ENVIRONMENTAL... Provisions § 80.8 Sampling methods for gasoline, diesel fuel, fuel additives, and renewable fuels. The..., blendstocks, fuel additives and renewable fuels for purposes of determining compliance with the...

  1. 40 CFR Appendix C to Part 50 - Measurement Principle and Calibration Procedure for the Measurement of Carbon Monoxide in the...

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... measured CO concentration. 2.2 The measurement system is calibrated by referencing the analyzer's CO... Additional information on calibration may be found in Section 12 of reference 1. 4.2 Apparatus. The major components and typical configurations of the calibration systems for the two calibration methods are shown...

  2. 40 CFR Appendix C to Part 50 - Measurement Principle and Calibration Procedure for the Measurement of Carbon Monoxide in the...

    Code of Federal Regulations, 2013 CFR

    2013-07-01

    ... measured CO concentration. 2.2The measurement system is calibrated by referencing the analyzer's CO....3Additional information on calibration may be found in Section 12 of reference 1. 4.2Apparatus. The major components and typical configurations of the calibration systems for the two calibration methods are shown...

  3. 40 CFR Appendix C to Part 50 - Measurement Principle and Calibration Procedure for the Measurement of Carbon Monoxide in the...

    Code of Federal Regulations, 2012 CFR

    2012-07-01

    ... measured CO concentration. 2.2The measurement system is calibrated by referencing the analyzer's CO....3Additional information on calibration may be found in Section 12 of reference 1. 4.2Apparatus. The major components and typical configurations of the calibration systems for the two calibration methods are shown...

  4. A Full-Envelope Air Data Calibration and Three-Dimensional Wind Estimation Method Using Global Output-Error Optimization and Flight-Test Techniques

    NASA Technical Reports Server (NTRS)

    Taylor, Brian R.

    2012-01-01

    A novel, efficient air data calibration method is proposed for aircraft with limited envelopes. This method uses output-error optimization on three-dimensional inertial velocities to estimate calibration and wind parameters. Calibration parameters are based on assumed calibration models for static pressure, angle of attack, and flank angle. Estimated wind parameters are the north, east, and down components. The only assumptions needed for this method are that the inertial velocities and Euler angles are accurate, the calibration models are correct, and that the steady-state component of wind is constant throughout the maneuver. A two-minute maneuver was designed to excite the aircraft over the range of air data calibration parameters and de-correlate the angle-of-attack bias from the vertical component of wind. Simulation of the X-48B (The Boeing Company, Chicago, Illinois) aircraft was used to validate the method, ultimately using data derived from wind-tunnel testing to simulate the un-calibrated air data measurements. Results from the simulation were accurate and robust to turbulence levels comparable to those observed in flight. Future experiments are planned to evaluate the proposed air data calibration in a flight environment.

  5. Does Your Optical Particle Counter Measure What You Think it Does? Calibration and Refractive Index Correction Methods.

    NASA Astrophysics Data System (ADS)

    Rosenberg, Phil; Dean, Angela; Williams, Paul; Dorsey, James; Minikin, Andreas; Pickering, Martyn; Petzold, Andreas

    2013-04-01

    Optical Particle Counters (OPCs) are the de-facto standard for in-situ measurements of airborne aerosol size distributions and small cloud particles over a wide size range. This is particularly the case on airborne platforms where fast response is important. OPCs measure scattered light from individual particles and generally bin particles according to the measured peak amount of light scattered (the OPC's response). Most manufacturers provide a table along with their instrument which indicates the particle diameters which represent the edges of each bin. It is important to correct the particle size reported by OPCs for the refractive index of the particles being measured, which is often not the same as for those used during calibration. However, the OPC's response is not a monotonic function of particle diameter and obvious problems occur when refractive index corrections are attempted, but multiple diameters correspond to the same OPC response. Here we recommend that OPCs are calibrated in terms of particle scattering cross section as this is a monotonic (usually linear) function of an OPC's response. We present a method for converting a bin's boundaries in terms of scattering cross section into a bin centre and bin width in terms of diameter for any aerosol species for which the scattering properties are known. The relationship between diameter and scattering cross section can be arbitrarily complex and does not need to be monotonic; it can be based on Mie-Lorenz theory or any other scattering theory. Software has been provided on the Sourceforge open source repository for scientific users to implement such methods in their own measurement and calibration routines. As a case study data is presented showing data from Passive Cavity Aerosol Spectrometer Probe (PCASP) and a Cloud Droplet Probe (CDP) calibrated using polystyrene latex spheres and glass beads before being deployed as part of the Fennec project to measure airborne dust in the inaccessible regions of

  6. Calibration and Validation of The Soil Water Balance Model Wave For Forest Stands In Flanders: 2. Methods and Preliminary Results

    NASA Astrophysics Data System (ADS)

    Verstraeten, W. W.; Minnaert, M.; Meiresonne, L.; van Slycken, J.; Lust, N.; Muys, B.; Feyen, J.

    This research, stimulated by government policy, is conducted in de framework of the reforestation of agricultural land and its impact on hydrological basins. The integra- tion of the ongoing research efforts in Flanders offers a unique opportunity to calibrate and validate the water balance module of the WAVE-model (Water and Agrochemicals in soil, crop and Vadose Environment; Van Clooster et al., 1994) for the most impor- tant soil and forest types. Hence, the scientific objectives of this project are multiple. This poster will show the first results of the calibrated and validated WAVE-model and the simulated evapotranspiration for forest stands and agricultural land use. Fur- thermore comparison will be made with sap-flow measurements using the Heat Field Deformation Method (Cermàk and Nadezhdina, 1998). The calibration and validation procedure consist of a sensitivity analysis of the model parameters, model calibration using simulated and measured time series of soil water content at different depths of the soil profile (March 2000-August 2001) and multi-site validation (using the cali- brated parameters from one site on other experimental sites). The crop factor (Kc), the saturated hydraulic conductivity (Ksat) and the root distribution function are the major calibration parameters. The Root Mean Square Error (RMSE), Model efficiency (ME) and Coefficient of Determination (CD) are the simulation statistics implemented during the calibration step. The preliminary results show time series of soil water con- tent (in cm3.cm-3 for the whole soil profile) with a RMSE of 2.97 % (optimal as small as possible), a ME of 0.58 (optimal 1.00) and a CD of 1.01 (optimal 1.00) for a beech stand at a validation site. In case of the popular stand the actual transpiration (Tact), simulated with WAVE is overestimated compared with the sap-flow measure- ments (June-September 2000). The simulation results show more peaks and fall to zero when heavily daily rain showers occur

  7. Mars Albedo Measurement in the Near IR Range for Additional Calibration of the TIRVIM Instrument of the ExoMars-2016 Mission

    NASA Astrophysics Data System (ADS)

    Maslov, I. A.; Shenavrin, V. I.; Grigoriev, A. V.; Moshkin, B. E.; Shakun, A. V.

    2017-01-01

    Results of ground-based measurements of the Mars albedo in the spectral range 1-5 μm, which were held in the days close to the session of measurements from the Mars orbit by the Russian device TIRVIM, are presented. The obtained data can be used to refine the calibration of the instrument.

  8. Non-Contact Thrust Stand Calibration Method for Repetitively-Pulsed Electric Thrusters

    NASA Technical Reports Server (NTRS)

    Wong, Andrea R.; Toftul, Alexandra; Polzin, Kurt A.; Pearson, J. Boise

    2011-01-01

    A thrust stand calibration technique for use in testing repetitively-pulsed electric thrusters for in-space propulsion has been developed and tested using a modified hanging pendulum thrust stand. In the implementation of this technique, current pulses are applied to a solenoidal coil to produce a pulsed magnetic field that acts against the magnetic field produced by a permanent magnet mounted to the thrust stand pendulum arm. The force on the magnet is applied in this non-contact manner, with the entire pulsed force transferred to the pendulum arm through a piezoelectric force transducer to provide a time-accurate force measurement. Modeling of the pendulum arm dynamics reveals that after an initial transient in thrust stand motion the quasisteady average deflection of the thrust stand arm away from the unforced or zero position can be related to the average applied force through a simple linear Hooke s law relationship. Modeling demonstrates that this technique is universally applicable except when the pulsing period is increased to the point where it approaches the period of natural thrust stand motion. Calibration data were obtained using a modified hanging pendulum thrust stand previously used for steady-state thrust measurements. Data were obtained for varying impulse bit at constant pulse frequency and for varying pulse frequency. The two data sets exhibit excellent quantitative agreement with each other as the constant relating average deflection and average thrust match within the errors on the linear regression curve fit of the data. Quantitatively, the error on the calibration coefficient is roughly 1% of the coefficient value.

  9. Regionalisation of a distributed method for flood quantiles estimation: Revaluation of local calibration hypothesis to enhance the spatial structure of the optimised parameter

    NASA Astrophysics Data System (ADS)

    Odry, Jean; Arnaud, Patrick

    2016-04-01

    The SHYREG method (Aubert et al., 2014) associates a stochastic rainfall generator and a rainfall-runoff model to produce rainfall and flood quantiles on a 1 km2 mesh covering the whole French territory. The rainfall generator is based on the description of rainy events by descriptive variables following probability distributions and is characterised by a high stability. This stochastic generator is fully regionalised, and the rainfall-runoff transformation is calibrated with a single parameter. Thanks to the stability of the approach, calibration can be performed against only flood quantiles associated with observated frequencies which can be extracted from relatively short time series. The aggregation of SHYREG flood quantiles to the catchment scale is performed using an areal reduction factor technique unique on the whole territory. Past studies demonstrated the accuracy of SHYREG flood quantiles estimation for catchments where flow data are available (Arnaud et al., 2015). Nevertheless, the parameter of the rainfall-runoff model is independently calibrated for each target catchment. As a consequence, this parameter plays a corrective role and compensates approximations and modelling errors which makes difficult to identify its proper spatial pattern. It is an inherent objective of the SHYREG approach to be completely regionalised in order to provide a complete and accurate flood quantiles database throughout France. Consequently, it appears necessary to identify the model configuration in which the calibrated parameter could be regionalised with acceptable performances. The revaluation of some of the method hypothesis is a necessary step before the regionalisation. Especially the inclusion or the modification of the spatial variability of imposed parameters (like production and transfer reservoir size, base flow addition and quantiles aggregation function) should lead to more realistic values of the only calibrated parameter. The objective of the work presented

  10. Development and analytical validation of a multivariate calibration method for determination of amoxicillin in suspension formulations by near infrared spectroscopy.

    PubMed

    Silva, Maurício A M; Ferreira, Marcus H; Braga, Jez W B; Sena, Marcelo M

    2012-01-30

    This paper proposes a new method for determination of amoxicillin in pharmaceutical suspension formulations, based on transflectance near infrared (NIR) measurements and partial least squares (PLS) multivariate calibration. A complete methodology was implemented for developing the proposed method, including an experimental design, data preprocessing by using multiple scatter correction (MSC) and outlier detection based on high values of leverage, and X and Y residuals. The best PLS model was obtained with seven latent variables in the range from 40.0 to 65.0 mg mL(-1) of amoxicillin, providing a root mean square error of prediction (RMSEP) of 1.6 mg mL(-1). The method was validated in accordance with Brazilian and international guidelines, through the estimate of figures of merit, such as linearity, precision, accuracy, robustness, selectivity, analytical sensitivity, limits of detection and quantitation, and bias. The results for determinations in four commercial pharmaceutical formulations were in agreement with the official high performance liquid chromatographic (HPLC) method at the 99% confidence level. A pseudo-univariate calibration curve was also obtained based on the net analyte signal (NAS). The proposed chemometric method presented the advantages of rapidity, simplicity, low cost, and no use of solvents, compared to the principal alternative methods based on HPLC.

  11. Simplified Vicarious Radiometric Calibration

    NASA Technical Reports Server (NTRS)

    Stanley, Thomas; Ryan, Robert; Holekamp, Kara; Pagnutti, Mary

    2010-01-01

    ground target areas having different reflectance values. The target areas can be natural or artificial and must be large enough to minimize adjacent-pixel contamination effects. The radiative coupling between the atmosphere and the terrain needs to be approximately the same for the two targets. This condition can be met for relatively uniform backgrounds when the distance between the targets is within a few hundred meters. For each target area, the radiance leaving the ground in the direction of the satellite is measured with a radiometrically calibrated spectroradiometer. Using the radiance measurements from the two targets, atmospheric adjacency and atmospheric scattering effects can be subtracted, thereby eliminating many assumptions about the atmosphere and the radiative interaction between the atmosphere and the terrain. In addition, the radiometrically calibrated spectroradiometer can be used with a known reflectance target to estimate atmospheric transmission and diffuse- to-global ratios without the need for ancillary sun photometers. Several comparisons between the simplified method and traditional techniques were found to agree within a few percent. Hence, the simplified method reduces the overall complexity of performing vicarious calibrations and can serve as a method for validating traditional radiative transfer models

  12. Toward a universal carbonate clumped isotope calibration: Diverse synthesis and preparatory methods suggest a single temperature relationship

    NASA Astrophysics Data System (ADS)

    Kelson, Julia R.; Huntington, Katharine W.; Schauer, Andrew J.; Saenger, Casey; Lechler, Alex R.

    2017-01-01

    Carbonate clumped isotope (Δ47) thermometry has been applied to a wide range of problems in earth, ocean and biological sciences over the last decade, but is still plagued by discrepancies among empirical calibrations that show a range of Δ47-temperature sensitivities. The most commonly suggested causes of these discrepancies are the method of mineral precipitation and analytical differences, including the temperature of phosphoric acid used to digest carbonates. However, these mechanisms have yet to be tested in a consistent analytical setting, which makes it difficult to isolate the cause(s) of discrepancies and to evaluate which synthetic calibration is most appropriate for natural samples. Here, we systematically explore the impact of synthetic carbonate precipitation by replicating precipitation experiments of previous workers under a constant analytical setting. We (1) precipitate 56 synthetic carbonates at temperatures of 4-85 °C using different procedures to degas CO2, with and without the use of the enzyme carbonic anhydrase (CA) to promote rapid dissolved inorganic carbon (DIC) equilibration; (2) digest samples in phosphoric acid at both 90 °C and 25 °C; and (3) hold constant all analytical methods including acid preparation, CO2 purification, and mass spectrometry; and (4) reduce our data with 17O corrections that are appropriate for our samples. We find that the CO2 degassing method does not influence Δ47 values of these synthetic carbonates, and therefore probably only influences natural samples with very rapid degassing rates, like speleothems that precipitate out of drip solution with high pCO2. CA in solution does not influence Δ47 values in this work, suggesting that disequilibrium in the DIC pool is negligible. We also find the Δ47 values of samples reacted in 25 and 90 °C acid are within error of each other (once corrected with a constant acid fractionation factor). Taken together, our results show that the Δ47-temperature relationship

  13. Transconvolution and the virtual positron emission tomograph-A new method for cross calibration in quantitative PET/CT imaging

    SciTech Connect

    Prenosil, George A.; Weitzel, Thilo; Hentschel, Michael; Klaeser, Bernd; Krause, Thomas

    2013-06-15

    with a Gaussian distribution were introduced. Furthermore, simulation of a virtual PET system provided a standard imaging system with clearly defined properties to which the real PET systems were to be matched. A Hann window served as the modulation transfer function for the virtual PET. The Hann's apodization properties suppressed high spatial frequencies above a certain critical frequency, thereby fulfilling the above-mentioned boundary conditions. The determined point spread functions were subsequently used by the novel Transconvolution algorithm to match different PET/CT systems onto the virtual PET system. Finally, the theoretically elaborated Transconvolution method was validated transforming phantom images acquired on two different PET systems to nearly identical data sets, as they would be imaged by the virtual PET system. Results: The proposed Transconvolution method matched different PET/CT-systems for an improved and reproducible determination of a normalized activity concentration. The highest difference in measured activity concentration between the two different PET systems of 18.2% was found in spheres of 2 ml volume. Transconvolution reduced this difference down to 1.6%. In addition to reestablishing comparability the new method with its parameterization of point spread functions allowed a full characterization of imaging properties of the examined tomographs. Conclusions: By matching different tomographs to a virtual standardized imaging system, Transconvolution opens a new comprehensive method for cross calibration in quantitative PET imaging. The use of a virtual PET system restores comparability between data sets from different PET systems by exerting a common, reproducible, and defined partial volume effect.

  14. Determination of gossypol content in cottonseeds by near infrared spectroscopy based on Monte Carlo uninformative variable elimination and nonlinear calibration methods.

    PubMed

    Li, Cheng; Zhao, Tianlun; Li, Cong; Mei, Lei; Yu, En; Dong, Yating; Chen, Jinhong; Zhu, Shuijin

    2017-04-15

    Near infrared (NIR) spectroscopy combined with Monte Carlo uninformative variable elimination (MC-UVE) and nonlinear calibration methods employed to determine gossypol content in cottonseeds were investigated. The reference method was performed by high performance liquid chromatography coupled to an ultraviolet detector (HPLC-UV). MC-UVE was employed to extract the effective information from the full NIR spectra. Nonlinear calibration methods were applied to establish the models compared with the linear method. The optimal model for gossypol content was obtained by MC-UVE-WLS-SVM, with root mean squares error of prediction (RMSEP) of 0.0422, coefficient of determination (R(2)) of 0.9331, and residual predictive deviation (RPD) of 3.8374, respectively, which was accurate and robust enough to substitute for traditional gossypol measurements. The nonlinear methods performed more reliable than linear method during the development of calibration models. Furthermore, MC-UVE could provide better and simpler calibration models than full spectra.

  15. Impacts of depolarization calibration methods on cloud phase interpretation at Eureka during 2013 and 2014 CRL lidar measurement campaigns

    NASA Astrophysics Data System (ADS)

    McCullough, E. M.; Perro, C. W.; Gamage, S. M.; Hopper, J.; Sica, R. J.; Duck, T.; Walker, K. A.; Drummond, J. R.

    2014-12-01

    The radiative behaviour of clouds is dependent on cloud particle phase. Water droplets can exist in temperatures well below 0° C for extended periods. Lidar depolarization measurements allow liquid and solid states to be differentiated in individual clouds at high spatial-temporal resolution. The 2012, 2013 and 2014 Canadian Arctic ACE Validation Campaigns in Eureka, Nunavut, Canada (80°N, 86°W) provided an opportunity to make extensive depolarization measurements using the CANDAC Rayleigh-Mie-Raman Lidar (CRL) in the troposphere.To date, most calibration methods in the literature are applicable to lidars which do not have non-ideal polarizing optics upstream of the polarizing analyzers in the receiver. We demonstrate a more complete matrix algebra calibration of the CRL to take the extra optics from six upstream lidar channels into account.Differences in depolarization parameter from 2013 and 2014 measurements show the advantage of the more extensive calibration for this lidar compared to the simpler traditional approach. The largest differences are found for depolarization parameter values around d = 0.50 (corresponding to δ = 0.33). Depolarization ratio values of δ = 0.2 to 0.3 are generally taken to be the cutoff between interpretations of ice (higher δ) or water (lower δ), and many CRL measurements lie in this particularly diagnostic range. An uncertainty analysis becomes important when extending the depolarization parameters to interpretation of the clouds in the atmosphere above Eureka.

  16. A Method of Calibrating Airspeed Installations on Airplanes at Transonic and Supersonic Speeds by the Use of Accelerometer and Attitude-Angle Measurements

    NASA Technical Reports Server (NTRS)

    Zalovick, John A; Lina, Lindsay J; Trant, James P , Jr

    1953-01-01

    A method is described for calibrating airspeed installation on airplanes at transonic and supersonic speeds in vertical-plane maneuvers in which use is made of measurements of normal and longitudinal accelerations and attitude angle. In this method all the required instrumentation is carried within the airplane. An analytical study of the effects of various sources of error on the accuracy of an airspeed calibration by the accelerometer method indicated that the required measurements can be made accurately enough to insure a satisfactory calibration.

  17. The aerospace imaging interferometer ALISEO: further improvements of calibration methods and assessment of interferometer response

    NASA Astrophysics Data System (ADS)

    Barducci, A.; Castagnoli, F.; Guzzi, D.; Marcoionni, P.; Pippi, I.

    2007-10-01

    ALISEO (Aerospace Leap-frog Imaging Stationary Interferometer for Earth Observation) belongs to the stationary interferometers representing a promising architecture for future Earth Observation (EO) sensors due to their simple optical layout. ALISEO has been selected by the Italian Space Agency as the principal payload for a new optical mission based on a micro-satellite (MIOsat). Payloads planned for MIOsat are an extensible telescope, a high-resolution panchromatic camera, a Mach-Zehnder MEMS interferometer, and ALISEO. MIOsat is expected to provide a platform with pointing capability for those advanced sensors. ALISEO operates in the common-path Sagnac configuration, and it does not employ any moving part to generate phase delay between the two rays. The sensor acquires the target images modulated by a pattern of autocorrelation functions: a fringe pattern that is fixed with respect to the instrument's field of view. The complete interferogram of each target location is retrieved introducing relative source-observer motion, which allows any image pixels to be observed under different phase delays. Recent laboratory measurements performed with ALISEO are described and discussed in this paper. In order to calibrate the optical path difference (OPD) of raw interferograms, a set of measurements have been carried out using a double planar diffuser system and several coloured He-Ne lasers. Standard reflectance tiles together doped with Holmium and Rare Earths have been used for validating the wavelength calibration of the instrument and proving the reliability of the reflectance retrieving procedure.

  18. An ensemble method based on uninformative variable elimination and mutual information for spectral multivariate calibration

    NASA Astrophysics Data System (ADS)

    Tan, Chao; Wang, Jinyue; Wu, Tong; Qin, Xin; Li, Menglong

    2010-12-01

    Based on the combination of uninformative variable elimination (UVE), bootstrap and mutual information (MI), a simple ensemble algorithm, named ESPLS, is proposed for spectral multivariate calibration (MVC). In ESPLS, those uninformative variables are first removed; and then a preparatory training set is produced by bootstrap, on which a MI spectrum of retained variables is calculated. The variables that exhibit higher MI than a defined threshold form a subspace on which a candidate partial least-squares (PLS) model is constructed. This process is repeated. After a number of candidate models are obtained, a small part of models is picked out to construct an ensemble model by simple/weighted average. Four near/mid-infrared (NIR/MIR) spectral datasets concerning the determination of six components are used to verify the proposed ESPLS. The results indicate that ESPLS is superior to UVEPLS and its combination with MI-based variable selection (SPLS) in terms of both the accuracy and robustness. Besides, from the perspective of end-users, ESPLS does not increase the complexity of a calibration when enhancing its performance.

  19. Clinical evaluation of a novel on-strip calibration method for blood glucose measurement.

    PubMed

    Noble, Michael; Rippeth, John; Edington, David; Rayman, Gerry; Brandon-Jones, Sarah; Hollowood, Zoe; Kew, Simon

    2014-07-01

    This study evaluated a novel technology for improving accuracy of self-monitoring of blood glucose (SMBG). The technology calibrates each and every test by measuring the response from a predetermined amount of glucose present in the sample chamber of each test strip. SMBG test strips were modified to include a lid coated with a fast dissolving formulation containing glucose. These test strips were characterized for hematocrit (Hct) and temperature induced error response to develop a calibration algorithm. The modified test strips were used in a clinical evaluation involving fingerstick blood samples from 160 subjects. Experiments involving Hct and temperature induced errors show that the technology generates a signal characteristic of the error conditions in any particular test, but independent of glucose concentration, allowing a correction algorithm to be derived. The approach substantially reduced Hct and temperature derived errors. Clinical evaluation using fingerstick blood directly applied to prototype strips showed the error (measured as MARD) was reduced from 11.1 to 5.9% by the on-strip correction approach and the number of outliers reduced by approximately 90%. This technology could improve the accuracy and precision of glucose monitoring systems and so reduce decision errors particularly in clinical situations where hematocrit and temperature may be significant confounders.

  20. Method and apparatus for calibrating a display using an array of cameras

    NASA Technical Reports Server (NTRS)

    Johnson, Michael J. (Inventor); Chen, Chung-Jen (Inventor); Chandrasekhar, Rajesh (Inventor)

    2001-01-01

    The present invention overcomes many of the disadvantages of the prior art by providing a display that can be calibrated and re-calibrated with a minimal amount of manual intervention. To accomplish this, the present invention provides one or more cameras to capture an image that is projected on a display screen. In one embodiment, the one or more cameras are placed on the same side of the screen as the projectors. In another embodiment, an array of cameras is provided on either or both sides of the screen for capturing a number of adjacent and/or overlapping capture images of the screen. In either of these embodiments, the resulting capture images are processed to identify any non-desirable characteristics including any visible artifacts such as seams, bands, rings, etc. Once the non-desirable characteristics are identified, an appropriate transformation function is determined. The transformation function is used to pre-warp the input video signal to the display such that the non-desirable characteristics are reduced or eliminated from the display. The transformation function preferably compensates for spatial non-uniformity, color non-uniformity, luminance non-uniformity, and/or other visible artifacts.

  1. Means and method for calibrating a photon detector utilizing electron-photon coincidence

    NASA Technical Reports Server (NTRS)

    Srivastava, S. K. (Inventor)

    1984-01-01

    An arrangement for calibrating a photon detector particularly applicable for the ultraviolet and vacuum ultraviolet regions is based on electron photon coincidence utilizing crossed electron beam atom beam collisions. Atoms are excited by electrons which lose a known amount of energy and scatter with a known remaining energy, while the excited atoms emit photons of known radiation. Electrons of the known remaining energy are separated from other electrons and are counted. Photons emitted in a direction related to the particular direction of scattered electrons are detected to serve as a standard. Each of the electrons is used to initiate the measurements of a time interval which terminates with the arrival of a photon exciting the photon detector. Only the number of time intervals related to the coincidence correlation and of electrons scattered in the particular direction with the known remaining energy and photons of a particular radiation level emitted due to the collisions of such scattered electrons are counted. The detector calibration is related to the number of counted electrons and photons.

  2. Standard addition method for the determination of pharmaceutical residues in drinking water by SPE-LC-MS/MS.

    PubMed

    Cimetiere, Nicolas; Soutrel, Isabelle; Lemasle, Marguerite; Laplanche, Alain; Crocq, André

    2013-01-01

    The study of the occurrence and fate of pharmaceutical compounds in drinking or waste water processes has become very popular in recent years. Liquid chromatography with tandem mass spectrometry is a powerful analytical tool often used to determine pharmaceutical residues at trace level in water. However, many steps may disrupt the analytical procedure and bias the results. A list of 27 environmentally relevant molecules, including various therapeutic classes and (cardiovascular, veterinary and human antibiotics, neuroleptics, non-steroidal anti-inflammatory drugs, hormones and other miscellaneous pharmaceutical compounds), was selected. In this work, a method was developed using ultra performance liquid chromatography coupled to tandem mass spectrometry (UPLC-MS/MS) and solid-phase extraction to determine the concentration of the 27 targeted pharmaceutical compounds at the nanogram per litre level. The matrix effect was evaluated from water sampled at different treatment stages. Conventional methods with external calibration and internal standard correction were compared with the standard addition method (SAM). An accurate determination of pharmaceutical compounds in drinking water was obtained by the SAM associated with UPLC-MS/MS. The developed method was used to evaluate the occurrence and fate of pharmaceutical compounds in some drinking water treatment plants in the west of France.

  3. Isotope-ratio-monitoring gas chromatography-mass spectrometry: methods for isotopic calibration

    NASA Technical Reports Server (NTRS)

    Merritt, D. A.; Brand, W. A.; Hayes, J. M.

    1994-01-01

    In trial analyses of a series of n-alkanes, precise determinations of 13C contents were based on isotopic standards introduced by five different techniques and results were compared. Specifically, organic-compound standards were coinjected with the analytes and carried through chromatography and combustion with them; or CO2 was supplied from a conventional inlet and mixed with the analyte in the ion source, or CO2 was supplied from an auxiliary mixing volume and transmitted to the source without interruption of the analyte stream. Additionally, two techniques were investigated in which the analyte stream was diverted and CO2 standards were placed on a near-zero background. All methods provided accurate results. Where applicable, methods not involving interruption of the analyte stream provided the highest performance (sigma = 0.00006 at.% 13C or 0.06% for 250 pmol C as CO2 reaching the ion source), but great care was required. Techniques involving diversion of the analyte stream were immune to interference from coeluting sample components and still provided high precision (0.0001 < or = sigma < or = 0.0002 at.% or 0.1 < or = sigma < or = 0.2%).

  4. A method for estimating spikelet number per panicle: Integrating image analysis and a 5-point calibration model.

    PubMed

    Zhao, Sanqin; Gu, Jiabing; Zhao, Youyong; Hassan, Muhammad; Li, Yinian; Ding, Weimin

    2015-11-06

    Spikelet number per panicle (SNPP) is one of the most important yield components used to estimate rice yields. The use of high-throughput quantitative image analysis methods for understanding the diversity of the panicle has increased rapidly. However, it is difficult to simultaneously extract panicle branch and spikelet/grain information from images at the same resolution due to the different scales of these traits. To use a lower resolution and meet the accuracy requirement, we proposed an interdisciplinary method that integrated image analysis and a 5-point calibration model to rapidly estimate SNPP. First, a linear relationship model between the total length of the primary branch (TLPB) and the SNPP was established based on the physiological characteristics of the panicle. Second, the TLPB and area (the primary branch region) traits were rapidly extracted by developing image analysis algorithm. Finally, a 5-point calibration method was adopted to improve the universality of the model. The number of panicle samples that the error of the SNPP estimates was less than 10% was greater than 90% by the proposed method. The estimation accuracy was consistent with the accuracy determined using manual measurements. The proposed method uses available concepts and techniques for automated estimations of rice yield information.

  5. An Automated Method of MFRSR Calibration for Aerosol Optical Depth Analysis with Application to an Asian Dust Outbreak over the United States.

    NASA Astrophysics Data System (ADS)

    Augustine, John A.; Cornwall, Christopher R.; Hodges, Gary B.; Long, Charles N.; Medina, Carlos I.; Deluisi, John J.

    2003-02-01

    Over the past decade, networks of Multifilter Rotating Shadowband Radiometers (MFRSR) and automated sun photometers have been established in the United States to monitor aerosol properties. The MFRSR alternately measures diffuse and global irradiance in six narrow spectral bands and a broadband channel of the solar spectrum, from which the direct normal component for each may be inferred. Its 500-nm channel mimics sun photometer measurements and thus is a source of aerosol optical depth information. Automatic data reduction methods are needed because of the high volume of data produced by the MFRSR. In addition, these instruments are often not calibrated for absolute irradiance and must be periodically calibrated for optical depth analysis using the Langley method. This process involves extrapolation to the signal the MFRSR would measure at the top of the atmosphere (I0). Here, an automated clear-sky identification algorithm is used