Science.gov

Sample records for addition calibration method

  1. In Situ Determination of Trace Elements in Fish Otoliths by Laser Ablation Double Focusing Sector Field Inductively Coupled Plasma Mass Spectrometry Using a Solution Standard Addition Calibration Method

    NASA Astrophysics Data System (ADS)

    Chen, Z.; Jones, C. M.

    2002-05-01

    Microchemistry of fish otoliths (fish ear bones) is a very useful tool for monitoring aquatic environments and fish migration. However, determination of the elemental composition in fish otolith by ICP-MS has been limited to either analysis of dissolved sample solution or measurement of limited number of trace elements by laser ablation (LA)- ICP-MS due to low sensitivity, lack of available calibration standards, and complexity of polyatomic molecular interference. In this study, a method was developed for in situ determination of trace elements in fish otoliths by laser ablation double focusing sector field ultra high sensitivity Finnigan Element 2 ICP-MS using a solution standard addition calibration method. Due to the lack of matrix-match solid calibration standards, sixteen trace elements (Na, Mg, P, Cr, Mn, Fe, Ni, Cu, Rb, Sr, Y, Cd, La, Ba, Pb and U) were determined using a solution standard calibration with Ca as an internal standard. Flexibility, easy preparation and stable signals are the advantages of using solution calibration standards. In order to resolve polyatomic molecular interferences, medium resolution (M/delta M > 4000) was used for some elements (Na, Mg, P, Cr, Mn, Fe, Ni, and Cu). Both external calibration and standard addition quantification strategies are compared and discussed. Precision, accuracy, and limits of detection are presented.

  2. Inspection system calibration methods

    DOEpatents

    Deason, Vance A.; Telschow, Kenneth L.

    2004-12-28

    An inspection system calibration method includes producing two sideband signals of a first wavefront; interfering the two sideband signals in a photorefractive material, producing an output signal therefrom having a frequency and a magnitude; and producing a phase modulated operational signal having a frequency different from the output signal frequency, a magnitude, and a phase modulation amplitude. The method includes determining a ratio of the operational signal magnitude to the output signal magnitude, determining a ratio of a 1st order Bessel function of the operational signal phase modulation amplitude to a 0th order Bessel function of the operational signal phase modulation amplitude, and comparing the magnitude ratio to the Bessel function ratio.

  3. Dilution standard addition calibration: A practical calibration strategy for multiresidue organic compounds determination.

    PubMed

    Martins, Manoel L; Rizzetti, Tiele M; Kemmerich, Magali; Saibt, Nathália; Prestes, Osmar D; Adaime, Martha B; Zanella, Renato

    2016-08-19

    Among calibration approaches for organic compounds determination in complex matrices, external calibration, based in solutions of the analytes in solvent or in blank matrix extracts, is the most applied approach. Although matrix matched calibration (MMC) can compensates the matrix effects, it does not compensate low recovery results. In this way, standard addition (SA) and procedural standard calibration (PSC) are usual alternatives, despite they generate more sample and/or matrix blanks consumption need, extra sample preparations and higher time and costs. Thus, the goal of this work was to establish a fast and efficient calibration approach, the diluted standard addition calibration (DSAC), based on successive dilutions of a spiked blank sample. In order to evaluate the proposed approach, solvent calibration (SC), MMC, PSC and DSAC were applied to evaluate recovery results of grape blank samples spiked with 66 pesticides. Samples were extracted with the acetate QuEChERS method and the compounds determined by ultra-high performance liquid chromatography tandem mass spectrometry (UHPLC-MS/MS). Results indicated that low recovery results for some pesticides were compensated by both PSC and DSAC approaches. Considering recoveries from 70 to 120% with RSD <20% as adequate, DSAC presented 83%, 98% and 100% of compounds meeting this criteria for the spiking levels 10, 50 and 100μgkg(-1), respectively. PSC presented same results (83%, 98% and 100%), better than those obtained by MMC (79%, 95% and 97%) and by SC (62%, 70% and 79%). The DSAC strategy showed to be suitable for calibration of multiresidue determination methods, producing adequate results in terms of trueness and is easier and faster to perform than other approaches. PMID:27432791

  4. Method for calibrating mass spectrometers

    DOEpatents

    Anderson, Gordon A [Benton City, WA; Brands, Michael D [Richland, WA; Bruce, James E [Schwenksville, PA; Pasa-Tolic, Ljiljana [Richland, WA; Smith, Richard D [Richland, WA

    2002-12-24

    A method whereby a mass spectra generated by a mass spectrometer is calibrated by shifting the parameters used by the spectrometer to assign masses to the spectra in a manner which reconciles the signal of ions within the spectra having equal mass but differing charge states, or by reconciling ions having known differences in mass to relative values consistent with those known differences. In this manner, the mass spectrometer is calibrated without the need for standards while allowing the generation of a highly accurate mass spectra by the instrument.

  5. Calibration method for an omnidirectional multicamera system

    NASA Astrophysics Data System (ADS)

    Ikeda, Sei; Sato, Tomokazu; Yokoya, Naokazu

    2003-05-01

    Telepresence systems using an omnidirectional image sensor enable us to experience remote site. A omnidirectional multi-camera system is more useful to acquire outdoor scenes than a monocular camera system, because the multi-camera system can easily capture high-resolution omnidirectional images. However, exact calibration of the camera system is necessary to virtualize the real world accurately. In this paper, we describe a geometric and photometric camera calibration and a panorama movie generation method for the omnidirectional multi-camera system. In the geometric calibration, intrinsic and extrinsic parameters of each camera are estimated using a calibration board and a laser measurement system called total station. In the photometric calibration, the limb darkening and color balances among the cameras are corrected. The result of the calibration is used in the panorama movie generation. In experiments, we have actually calibrated the multi-camera system and have generated spherical panorama movies by using the estimated camera parameters. A telepresence system was prototyped in order to confirm that the panorama movie can be used for telepresence well. In addition, we have evaluated the discontinuity in generated panoramic images.

  6. Calibration method for spectroscopic systems

    DOEpatents

    Sandison, David R.

    1998-01-01

    Calibration spots of optically-characterized material placed in the field of view of a spectroscopic system allow calibration of the spectroscopic system. Response from the calibration spots is measured and used to calibrate for varying spectroscopic system operating parameters. The accurate calibration achieved allows quantitative spectroscopic analysis of responses taken at different times, different excitation conditions, and of different targets.

  7. Calibration method for spectroscopic systems

    DOEpatents

    Sandison, D.R.

    1998-11-17

    Calibration spots of optically-characterized material placed in the field of view of a spectroscopic system allow calibration of the spectroscopic system. Response from the calibration spots is measured and used to calibrate for varying spectroscopic system operating parameters. The accurate calibration achieved allows quantitative spectroscopic analysis of responses taken at different times, different excitation conditions, and of different targets. 3 figs.

  8. MODIS Radiometric Calibration Program, Methods and Results

    NASA Technical Reports Server (NTRS)

    Xiong, Xiaoxiong; Guenther, Bruce; Angal, Amit; Barnes, William; Salomonson, Vincent; Sun, Junqiang; Wenny, Brian

    2012-01-01

    As a key instrument for NASA s Earth Observing System (EOS), the Moderate Resolution Imaging Spectroradiometer (MODIS) has made significant contributions to the remote sensing community with its unprecedented amount of data products continuously generated from its observations and freely distributed to users worldwide. MODIS observations, covering spectral regions from visible (VIS) to long-wave infrared (LWIR), have enabled a broad range of research activities and applications for studies of the earth s interactive system of land, oceans, and atmosphere. In addition to extensive pre-launch measurements, developed to characterize sensor performance, MODIS carries a set of on-board calibrators (OBC) that can be used to track on-orbit changes of various sensor characteristics. Most importantly, dedicated and continuous calibration efforts have been made to maintain sensor data quality. This paper provides an overview of the MODIS calibration program, on-orbit calibration activities, methods, and performance. Key calibration results and lessons learned from the MODIS calibration effort are also presented in this paper.

  9. A proposed standard method for polarimetric calibration and calibration verification

    NASA Astrophysics Data System (ADS)

    Persons, Christopher M.; Jones, Michael W.; Farlow, Craig A.; Morell, L. Denise; Gulley, Michael G.; Spradley, Kevin D.

    2007-09-01

    Accurate calibration of polarimetric sensors is critical to reducing and analyzing phenomenology data, producing uniform polarimetric imagery for deployable sensors, and ensuring predictable performance of polarimetric algorithms. It is desirable to develop a standard calibration method, including verification reporting, in order to increase credibility with customers and foster communication and understanding within the polarimetric community. This paper seeks to facilitate discussions within the community on arriving at such standards. Both the calibration and verification methods presented here are performed easily with common polarimetric equipment, and are applicable to visible and infrared systems with either partial Stokes or full Stokes sensitivity. The calibration procedure has been used on infrared and visible polarimetric imagers over a six year period, and resulting imagery has been presented previously at conferences and workshops. The proposed calibration method involves the familiar calculation of the polarimetric data reduction matrix by measuring the polarimeter's response to a set of input Stokes vectors. With this method, however, linear combinations of Stokes vectors are used to generate highly accurate input states. This allows the direct measurement of all system effects, in contrast with fitting modeled calibration parameters to measured data. This direct measurement of the data reduction matrix allows higher order effects that are difficult to model to be discovered and corrected for in calibration. This paper begins with a detailed tutorial on the proposed calibration and verification reporting methods. Example results are then presented for a LWIR rotating half-wave retarder polarimeter.

  10. A practical method for sensor absolute calibration.

    PubMed

    Meisenholder, G W

    1966-04-01

    This paper describes a method of performing sensor calibrations using an NBS standard of spectral irradiance. The method shown, among others, was used for calibration of the Mariner IV Canopus sensor. Agreement of inflight response to preflight calibrations performed by this technique has been found to be well within 10%. PMID:20048890

  11. A Comparison of Two Balance Calibration Model Building Methods

    NASA Technical Reports Server (NTRS)

    DeLoach, Richard; Ulbrich, Norbert

    2007-01-01

    Simulated strain-gage balance calibration data is used to compare the accuracy of two balance calibration model building methods for different noise environments and calibration experiment designs. The first building method obtains a math model for the analysis of balance calibration data after applying a candidate math model search algorithm to the calibration data set. The second building method uses stepwise regression analysis in order to construct a model for the analysis. Four balance calibration data sets were simulated in order to compare the accuracy of the two math model building methods. The simulated data sets were prepared using the traditional One Factor At a Time (OFAT) technique and the Modern Design of Experiments (MDOE) approach. Random and systematic errors were introduced in the simulated calibration data sets in order to study their influence on the math model building methods. Residuals of the fitted calibration responses and other statistical metrics were compared in order to evaluate the calibration models developed with different combinations of noise environment, experiment design, and model building method. Overall, predicted math models and residuals of both math model building methods show very good agreement. Significant differences in model quality were attributable to noise environment, experiment design, and their interaction. Generally, the addition of systematic error significantly degraded the quality of calibration models developed from OFAT data by either method, but MDOE experiment designs were more robust with respect to the introduction of a systematic component of the unexplained variance.

  12. Study on self-calibration angle encoder using simulation method

    NASA Astrophysics Data System (ADS)

    Wang, Yan; Xue, Zi; Huang, Yao; Wang, Xiaona

    2016-01-01

    The angle measurement technology is very important in precision manufacture, optical industry, aerospace, aviation and navigation, etc. Further, the angle encoder, which uses concept `subdivision of full circle (2π rad=360°)' and transforms the angle into number of electronic pulse, is the most common instrument for angle measurement. To improve the accuracy of the angle encoder, a novel self-calibration method was proposed that enables the angle encoder to calibrate itself without angle reference. An angle deviation curve among 0° to 360° was simulated with equal weights Fourier components for the study of the self-calibration method. In addition, a self-calibration algorithm was used in the process of this deviation curve. The simulation result shows the relationship between the arrangement of multi-reading heads and the Fourier components distribution of angle encoder deviation curve. Besides, an actual self-calibration angle encoder was calibrated by polygon angle standard in national institute of metrology, China. The experiment result indicates the actual self-calibration effect on the Fourier components distribution of angle encoder deviation curve. In the end, the comparison, which is between the simulation self-calibration result and the experiment self-calibration result, reflects good consistency and proves the reliability of the self-calibration angle encoder.

  13. Automatic alignment method for calibration of hydrometers

    NASA Astrophysics Data System (ADS)

    Lee, Y. J.; Chang, K. H.; Chon, J. C.; Oh, C. Y.

    2004-04-01

    This paper presents a new method to automatically align specific scale-marks for the calibration of hydrometers. A hydrometer calibration system adopting the new method consists of a vision system, a stepping motor, and software to control the system. The vision system is composed of a CCD camera and a frame grabber, and is used to acquire images. The stepping motor moves the camera, which is attached to the vessel containing a reference liquid, along the hydrometer. The operating program has two main functions: to process images from the camera to find the position of the horizontal plane and to control the stepping motor for the alignment of the horizontal plane with a particular scale-mark. Any system adopting this automatic alignment method is a convenient and precise means of calibrating a hydrometer. The performance of the proposed method is illustrated by comparing the calibration results using the automatic alignment method with those obtained using the manual method.

  14. Iterative optimization calibration method for stereo deflectometry.

    PubMed

    Ren, Hongyu; Gao, Feng; Jiang, Xiangqian

    2015-08-24

    An accurate system calibration method is presented in this paper to calibrate stereo deflectometry. A corresponding iterative optimization algorithm is also proposed to improve the system calibration accuracy. This merges CCD parameters and geometrical relation between CCDs and the LCD into one cost function. In this calibration technique, an optical flat acts as a reference mirror and simultaneously reflect sinusoidal fringe patterns into the two CCDs. The normal vector of the reference mirror is used as an intermediate variable to implement this iterative optimization algorithm until the root mean square of the reprojection errors converge to a minimum. The experiment demonstrates that this method can optimize all the calibration parameters and can effectively reduce reprojection error, which correspondingly improves the final reconstruction accuracy. PMID:26368180

  15. Two highly accurate methods for pitch calibration

    NASA Astrophysics Data System (ADS)

    Kniel, K.; Härtig, F.; Osawa, S.; Sato, O.

    2009-11-01

    Among profiles, helix and tooth thickness pitch is one of the most important parameters of an involute gear measurement evaluation. In principle, coordinate measuring machines (CMM) and CNC-controlled gear measuring machines as a variant of a CMM are suited for these kinds of gear measurements. Now the Japan National Institute of Advanced Industrial Science and Technology (NMIJ/AIST) and the German national metrology institute the Physikalisch-Technische Bundesanstalt (PTB) have each developed independently highly accurate pitch calibration methods applicable to CMM or gear measuring machines. Both calibration methods are based on the so-called closure technique which allows the separation of the systematic errors of the measurement device and the errors of the gear. For the verification of both calibration methods, NMIJ/AIST and PTB performed measurements on a specially designed pitch artifact. The comparison of the results shows that both methods can be used for highly accurate calibrations of pitch standards.

  16. Design and calibration of zero-additional-phase SPIDER

    SciTech Connect

    Baum, Peter; Riedle, Eberhard

    2005-09-01

    Zero-additional-phase spectral phase interferometry for direct electric field reconstruction (ZAP-SPIDER) is a novel technique for measuring the temporal shape and phase of ultrashort optical pulses directly at the interaction point of a spectroscopic experiment. The scheme is suitable for an extremely wide wavelength region from the ultraviolet to the near infrared. We present a comprehensive description of the experimental setup and design guidelines to effectively apply the technique to various wavelengths and pulse durations. The calibration of the setup and procedures to check the consistency of the measurement are discussed in detail. We show experimental data for various center wavelengths and pulse durations down to 7 fs to verify the applicability to a wide range of pulse parameters.

  17. Method of Calibrating a Force Balance

    NASA Technical Reports Server (NTRS)

    Parker, Peter A. (Inventor); Rhew, Ray D. (Inventor); Johnson, Thomas H. (Inventor); Landman, Drew (Inventor)

    2015-01-01

    A calibration system and method utilizes acceleration of a mass to generate a force on the mass. An expected value of the force is calculated based on the magnitude and acceleration of the mass. A fixture is utilized to mount the mass to a force balance, and the force balance is calibrated to provide a reading consistent with the expected force determined for a given acceleration. The acceleration can be varied to provide different expected forces, and the force balance can be calibrated for different applied forces. The acceleration may result from linear acceleration of the mass or rotational movement of the mass.

  18. [High Precision Spectral Calibration Method of Fourier Interferometric Spectrometer].

    PubMed

    Lin, Jun; Shao, Jun; Song, Chao-yu; Li, Yun-wei; Lei, Yu-fei

    2015-12-01

    The Fourier interferometric spectrometer (FIS) acquires the interference data information of the spectrum and during the spectrum data processing, a series of spectrum reconstruction will be performed on the interference information to obtain the final spectrum information data. The spectral calibration is the key step to spectrum reconstruction of FIS, which directly determines accuracy and availability of the spectrum results. This paper introduces the basic ideas and calibration accuracy about the spectral calibration for the FIS and puts forward a new spectral calibration method based on calculating the precise value of the total optical path difference (TOPD). The TOPD of FIS is difficult to be precisely measured, but it is the core and key to the spectral calibration. In order to calculate the precise TOPD, this paper proposes the idea how to traverse the TOPD and analyzes the spectrum drift. During the calibration, all the possible values of the TOPD participate in the spectrum reconstruction flow to carry out spectrum recovery and analysis. Ultimately the TOPD with the minimum spectrum drift will be achieved, namely solution value of the TOPD. This method can accurately resolve the TOPD of the FIS and then calibrate the spectrum with high accuracy. In addition, the paper introduces the detailed and complete spectral calibration flow and obtains the center wavelength value of every band and wavenumber resolution. Moreover, the paper designs the main parameters of the typical FIS and generates its simulation interference data. Using the above method to calibrate the simulation data, the analysis and verification of the spectral calibration results proves that the calibration precision of wavenumber resolution achieves 0.000 25 cm⁻¹ or above. PMID:26964245

  19. New Method of Calibrating IRT Models.

    ERIC Educational Resources Information Center

    Jiang, Hai; Tang, K. Linda

    This discussion of new methods for calibrating item response theory (IRT) models looks into new optimization procedures, such as the Genetic Algorithm (GA) to improve on the use of the Newton-Raphson procedure. The advantages of using a global optimization procedure like GA is that this kind of procedure is not easily affected by local optima and…

  20. Automatic calibration method for plenoptic camera

    NASA Astrophysics Data System (ADS)

    Luan, Yinsen; He, Xing; Xu, Bing; Yang, Ping; Tang, Guomao

    2016-04-01

    An automatic calibration method is proposed for a microlens-based plenoptic camera. First, all microlens images on the white image are searched and recognized automatically based on digital morphology. Then, the center points of microlens images are rearranged according to their relative position relationships. Consequently, the microlens images are located, i.e., the plenoptic camera is calibrated without the prior knowledge of camera parameters. Furthermore, this method is appropriate for all types of microlens-based plenoptic cameras, even the multifocus plenoptic camera, the plenoptic camera with arbitrarily arranged microlenses, or the plenoptic camera with different sizes of microlenses. Finally, we verify our method by the raw data of Lytro. The experiments show that our method has higher intelligence than the methods published before.

  1. Calibration method for radiometric and wavelength calibration of a spectrometer

    NASA Astrophysics Data System (ADS)

    Granger, Edward M.

    1998-12-01

    A new calibration target or Certified Reference Material (CRM) has been designed that uses violet, orange, green and cyan dyes ont cotton paper. This paper type was chosen because it has a relatively flat spectral response from 400 nm to 700 nm and good keeping properties. These specific dyes were chosen because the difference signal between the orange, cyan, green and purple dyes have certain characteristics that then a low the calibration of an instrument. The ratio between the difference readings is a direct function of the center wavelength of a given spectral band. Therefore, the radiometric and spectral calibration can be determined simultaneously from the physical properties of the reference materials.

  2. Development and validation of a dynamic range-extended LC-MS/MS multi-analyte method for 11 different postmortem matrices for redistribution studies applying solvent calibration and additional (13)C isotope monitoring.

    PubMed

    Staeheli, Sandra N; Poetzsch, Michael; Kraemer, Thomas; Steuer, Andrea E

    2015-11-01

    Postmortem redistribution (PMR) is one of numerous problems in postmortem toxicology making correct interpretation of measured drug concentrations difficult or even impossible. Time-dependent PMR in peripheral blood and especially in tissue samples is still under-explored. For further investigation, an easy applicable method for the simultaneous quantitation of over 80 forensically relevant compounds in 11 different postmortem matrices should be developed and validated overcoming the challenges of high inter-matrix and intra-matrix concentration variances. Biopsy samples (20 mg) or body fluids (20 μL) were spiked with an analyte mix and deuterated internal standards, extracted by liquid-liquid extraction, and analyzed by liquid chromatography tandem mass spectrometry (LC-MS/MS). For highest applicability, an easy solvent calibration was used. Furthermore, time-consuming dilution of high concentration samples showing detector saturation was circumvented by two overlapping calibration curves using (12)C isotope monitoring for low concentrations and (13)C isotopes for high concentration, respectively. The method was validated according to international guidelines with modifications. Matrix effects and extraction efficiency were strongly matrix and analyte dependent. In general, brain and adipose tissue produced the highest matrix effects, whereas cerebrospinal fluid showed the least matrix effects. Accuracy and precision results were rather matrix independent with some exceptions. Despite using an external solvent calibration, the accuracy requirements were fulfilled for 66 to 81 % of the 83 analytes. Depending on the matrix, 75-93 % of the analytes showed intra-day precisions at <20 %. (12)C and (13)C calibrations gave comparable results and proved to be a useful tool in expanding the dynamic range. PMID:26396081

  3. A new continuous calibration method for inductively coupled plasma spectrometry.

    PubMed

    Paredes, E; Maestre, S E; Todolí, J L

    2006-01-01

    A new calibration method was developed and applied to inductively coupled plasma atomic emission spectrometry. External calibration was performed as follows. A container was filled with a given volume of deionized (V(p)) water. Then a concentrated standard was introduced at a controlled rate (Q(e)) into the tank by means of a peristaltic pump. The resulting solution was stirred throughout the experiment. Simultaneously, the solution inside the tank was pumped from the vessel to the plasma at a given rate (Q(s)). The signal was continuously recorded. The variation of the concentration of the solution leaving the tank with time was determined by applying a basic equation of stirred tanks. The representation of the emission intensity versus the time and the further conversion of the time scale into a concentration scale gave rise to the calibration line. The best results in terms of linearity were achieved for V(p)=15 cm3, Q(e)=0.6-0.75 ml min(-1) and Q(s)=1-1.2 ml min(-1). Graphs with more than 40 standards were obtained within about 10 min. The results found were not statistically different from those afforded by the conventional calibration method. In addition, the new method was faster and supplied better linearity and precision than the conventional one. Another advantage of the stirred tank was that procedures such as dynamic calibration and standard additions could be easily and quickly applied, thus shortening the analysis time. A complete analysis following these procedures based on the measurement of 30 standards took about 5 min. Several synthetic as well as certified samples (i.e., bovine liver, mussel tissue and powdered milk) were analyzed with the stirred tank by applying four different calibration methodologies (i.e., external calibration, internal calibration, standard additions and a combination of internal standardization and standard additions), with the combination of internal standardization and standard additions being the method that provided

  4. Research on calibration method of relative infrared radiometer

    NASA Astrophysics Data System (ADS)

    Yang, Sen; Li, Chengwei

    2016-02-01

    The Relative Infrared Radiometer (RIR) is commonly used to measure the irradiance of the Infrared Target Simulator (ITS), and the calibration of the RIR is central for the measurement accuracy. RIR calibration is conventionally performed using the Radiance Based (RB) calibration method or Irradiance Based (IB) calibration method, and the relationship between the radiation of standard source and the response of RIR is determined by curve fitting. One limitation existing in the calibration of RIR is the undesirable calibration voltage fluctuation in single measurement or in the reproducibility measurement, which reduces the calibration reproducibility and irradiance measurement accuracy. To address this limitation, the Equivalent Blackbody Temperature Based (EBTB) calibration method is proposed for the calibration of RIR. The purpose of this study is to compare the proposed EBTB calibration method with conventional RB and IB calibration methods. The comparison and experiment results have shown that the EBTB calibration method is not only able to provide comparable correlation between radiation and response to other calibration methods (IB and RB) in the irradiance measurement but also reduces the influence of calibration voltage fluctuation on the irradiance measurement result, which improves the calibration reproducibility and irradiance measurement accuracy.

  5. Accurate wavelength calibration method for flat-field grating spectrometers.

    PubMed

    Du, Xuewei; Li, Chaoyang; Xu, Zhe; Wang, Qiuping

    2011-09-01

    A portable spectrometer prototype is built to study wavelength calibration for flat-field grating spectrometers. An accurate calibration method called parameter fitting is presented. Both optical and structural parameters of the spectrometer are included in the wavelength calibration model, which accurately describes the relationship between wavelength and pixel position. Along with higher calibration accuracy, the proposed calibration method can provide information about errors in the installation of the optical components, which will be helpful for spectrometer alignment. PMID:21929865

  6. A webcam photogrammetric method for robot calibration

    NASA Astrophysics Data System (ADS)

    Sargeant, Ben; Hosseininaveh, Ali A.; Erfani, Tohid; Robson, Stuart; Boehm, Jan

    2013-04-01

    This paper describes a strategy for accurate robot calibration using close range photogrammetry. A 5-DoF robot has been designed for placement of two web cameras relative to an object. To ensure correct camera positioning, the robot is calibrated using the following strategy. First, a Denavit-Hartenberg method is used to generate a general kinematic robot model. A set of reference frames are defined relative to each joint and each of the cameras, transformation matrices are then produced to represent change in position and orientation between frames in terms of joint positions and unknown parameters. The complete model is extracted by multiplying these matrices. Second, photogrammetry is used to estimate the postures of both cameras. A set of images are captured of a calibration fixture from different robot poses. The camera postures are then estimated using bundle adjustment. Third, the kinematic parameters are estimated using weighted least squares. For each pose a set of equations are extracted from the model and the unknown parameters are estimated in an iterative procedure. Finally these values are substituted back into the original model. This final model is tested using forward kinematics by comparing the model's predicted camera postures for given joint positions to the values obtained through photogrammetry. Inverse kinematics is performed using both least squares and particle swarm optimisation and these techniques are contrasted. Results demonstrate that this photogrammetry approach produces a reliable and accurate model of the robot that can be used with both least squares and particle swarm optimisation for robot control.

  7. Optimal Multicomponent Analysis Using the Generalized Standard Addition Method.

    ERIC Educational Resources Information Center

    Raymond, Margaret; And Others

    1983-01-01

    Describes an experiment on the simultaneous determination of chromium and magnesium by spectophotometry modified to include the Generalized Standard Addition Method computer program, a multivariate calibration method that provides optimal multicomponent analysis in the presence of interference and matrix effects. Provides instructions for…

  8. New methods of measuring and calibrating robots

    NASA Astrophysics Data System (ADS)

    Janocha, Hartmut; Diewald, Bernd

    1995-10-01

    ISO 9283 and RIA R15.05 define industrial robot parameters which are applied to compare the efficiency of different robots. Hitherto, however, no suitable measurement systems have been available. ICAROS is a system which combines photogrammetrical procedures with an inertial navigation system. For the first time, this combination allows the high-precision static and dynamic measurement of the position as well as of the orientation of the robot endeffector. Thus, not only the measuring data for the determination of all industrial robot parameters can be acquired. By integration of a new over-all-calibration procedure, ICAROS also allows the reduction of the absolute robot pose errors to the range of its repeatability. The integration of both system components as well as measurement and calibration results are presented in this paper, using a six-axes robot as example. A further approach also presented here takes into consideration not only the individual robot errors but also the tolerances of workpieces. This allows the adjustment of off-line programs of robots based on inexact or idealized CAD data in any pose. Thus the robot position which is defined relative to the workpiece to be processed, is achieved as required. This includes the possibility to transfer teached robot programs to other devices without additional expenditure. The adjustment is based on the measurement of the robot position using two miniaturized CCD cameras mounted near the endeffector which are carried along by the robot during the correction phase. In the area viewed by both cameras, the robot position is determined in relation to prominent geometry elements, e.g. lines or holes. The scheduled data to be compared therewith can either be calculated in modern off-line programming systems during robot programming, or they can be determined at the so-called master robot if a transfer of the robot program is desired.

  9. A dynamic calibration method for the pressure transducer

    NASA Astrophysics Data System (ADS)

    Wang, Zhongyu; Wang, Zhuoran; Li, Qiang

    2016-01-01

    Pressure transducer is widely used in the field of industry. A calibrated pressure transducer can increase the performance of precision instruments in the closed mechanical relationship. Calibration is the key to ensure the pressure transducer with a high precision and dynamic characteristic. Unfortunately, the current calibration method can usually be used in the laboratory with a good condition and only one pressure transducer can be calibrated at each time. Therefore the calibration efficiency is hard to meet the requirement of modern industry with high efficiency. A dynamic and fast calibration technology with a calibration device and a corresponding data processing method is proposed in this paper. Firstly, the pressure transducer to be calibrated is placed in the small cavity chamber. The calibration process only contains a single loop. The outputs of each calibrated transducer are recorded automatically by the control terminal. Secondly, LabView programming is used for the information acquisition and data processing. The performance of the repeatability and nonlinear indicators can be figured out directly. At last the pressure transducers are calibrated simultaneously in the experiment to verify the suggested calibration technology. The experimental result shows this method can be used to calibrate the pressure transducer in the practical engineering measurement.

  10. Method for calibration of plutonium NDA

    SciTech Connect

    Lemming, J.F.; Campbell, A.R.; Rodenburg, W.W.

    1980-01-01

    Calibration materials characterized by calorimetric assay can be a practical alternative to synthetic standards for the calibration of plutonium nondestructive assay. Calorimetric assay is an effective measurement system for the characterization because: it can give an absolute assay from first principles when the isotopic composition is known, it is insensitive to most matrix effects, and its traceability to international measurement systems has been demonstrated.

  11. Design Method and Calibration of Moulinet

    NASA Astrophysics Data System (ADS)

    Itoh, Hirokazu; Yamada, Hirokazu; Udagawa, Sinsuke

    The formula for obtaining the absorption horsepower of a Moulinet was rewritten, and the physical meaning of the constant in the formula was clarified. Based on this study, the design method of the Moulinet and the calibration method of the Moulinet that was performed after manufacture were verified experimentally. Consequently, the following was clarified; (1) If the propeller power coefficient was taken to be the proportionality constant, the absorption horsepower of the Moulinet was proportional to the cube of the revolution speed, and the fifth power of the Moulinet diameter. (2) If the Moulinet design was geometrically similar to the standard dimensions of the Aviation Technical Research Center's type-6 Moulinet, the proportionality constant C1 given in the reference could be used, and the absorption horsepower of the Moulinet was proportional to the cube of the revolution speed, the cube of the Moulinet diameter, and the side projection area of the Moulinet. (3) The proportionality constant C1 was proportional to the propeller power coefficient CP.

  12. Pulsed Electric Propulsion Thrust Stand Calibration Method

    NASA Technical Reports Server (NTRS)

    Wong, Andrea R.; Polzin, Kurt A.; Pearson, J. Boise

    2011-01-01

    The evaluation of the performance of any propulsion device requires the accurate measurement of thrust. While chemical rocket thrust is typically measured using a load cell, the low thrust levels associated with electric propulsion (EP) systems necessitate the use of much more sensitive measurement techniques. The design and development of electric propulsion thrust stands that employ a conventional hanging pendulum arm connected to a balance mechanism consisting of a secondary arm and variable linkage have been reported in recent publications by Polzin et al. These works focused on performing steady-state thrust measurements and employed a static analysis of the thrust stand response. In the present work, we present a calibration method and data that will permit pulsed thrust measurements using the Variable Amplitude Hanging Pendulum with Extended Range (VAHPER) thrust stand. Pulsed thrust measurements are challenging in general because the pulsed thrust (impulse bit) occurs over a short timescale (typically 1 micros to 1 millisecond) and cannot be resolved directly. Consequently, the imparted impulse bit must be inferred through observation of the change in thrust stand motion effected by the pulse. Pulsed thrust measurements have typically only consisted of single-shot operation. In the present work, we discuss repetition-rate pulsed thruster operation and describe a method to perform these measurements. The thrust stand response can be modeled as a spring-mass-damper system with a repetitive delta forcing function to represent the impulsive action of the thruster.

  13. Camera self-calibration method based on two vanishing points

    NASA Astrophysics Data System (ADS)

    Duan, Shaoli; Zang, Huaping; Xu, Mengmeng; Zhang, Xiaofang; Gong, Qiaoxia; Tian, Yongzhi; Liang, Erjun; Liu, Xiaomin

    2015-10-01

    Camera calibration is one of the indispensable processes to obtain 3D depth information from 2D images in the field of computer vision. Camera self-calibration is more convenient and flexible, especially in the application of large depth of fields, wide fields of view, and scene conversion, as well as other occasions like zooms. In this paper, a self-calibration method based on two vanishing points is proposed, the geometric characteristic of disappear points formed by two groups of orthogonal parallel lines is applied to camera self-calibration. By using the vectors' orthogonal properties of connection optical centers and the vanishing points, the constraint equations on the camera intrinsic parameters are established. By this method, four internal parameters of the camera can be solved though only four images taken from different viewpoints in a scene. Compared with the two other self-calibration methods with absolute quadric and calibration plate, the method based on two vanishing points does not require calibration objects, camera movement, the information on the size and location of parallel lines, without strict experimental equipment, and having convenient calibration process and simple algorithm. Compared with the experimental results of the method based on calibration plate, self-calibration method by using machine vision software Halcon, the practicability and effectiveness of the proposed method in this paper is verified.

  14. Calibration methods for rotating shadowband irradiometers and evaluation of calibration duration

    NASA Astrophysics Data System (ADS)

    Jessen, W.; Wilbert, S.; Nouri, B.; Geuder, N.; Fritz, H.

    2015-10-01

    Resource assessment for Concentrated Solar Power (CSP) needs accurate Direct Normal Irradiance (DNI) measurements. An option for such measurement campaigns are Rotating Shadowband Irradiometers (RSIs) with a thorough calibration. Calibration of RSIs and Si-sensors in general is complex because of the inhomogeneous spectral response of such sensors and incorporates the use of several correction functions. A calibration for a given atmospheric condition and air mass might not work well for a different condition. This paper covers procedures and requirements for two calibration methods for the calibration of Rotating Shadowband Irradiometers. The necessary duration of acquisition of test measurements is examined in regard to the site specific conditions at Plataforma Solar de Almeria (PSA) in Spain. Data sets of several long-term calibration periods from PSA are used to evaluate the deviation of results from calibrations with varying duration from the long-term result. The findings show that seasonal changes of environmental conditions are causing small but noticeable fluctuation of calibration results. Certain periods (i.e. November to January and April to May) show a higher likelihood of particularly adverse calibration results. These effects can partially be compensated by increasing the inclusions of measurements from outside these periods. Consequently, the duration of calibrations at PSA can now be selected depending on the time of the year in which measurements are commenced.

  15. Calibration of a catchment scale cosmic-ray probe network: A comparison of three parameterization methods

    NASA Astrophysics Data System (ADS)

    Baatz, R.; Bogena, H. R.; Hendricks Franssen, H.-J.; Huisman, J. A.; Qu, W.; Montzka, C.; Vereecken, H.

    2014-08-01

    The objective of this work was to assess the accuracy of soil water content determination from neutron flux measured by cosmic-ray probes under humid climate conditions. Ten cosmic-ray probes were set up in the Rur catchment located in western Germany, and calibrated by gravimetric soil sampling campaigns. Aboveground biomass was estimated at the sites to investigate the role of vegetation cover on the neutron flux and the calibration procedure. Three parameterization methods were used to generate site-specific neutron flux - soil water content calibration curves: (i) the N0-method, (ii) the hydrogen molar fraction method (hmf-method), and (iii) the COSMIC-method. At five locations, calibration measurements were repeated to evaluate site-specific calibration parameters obtained in two different sampling campaigns. At two locations, soil water content determined by cosmic-ray probes was evaluated with horizontally and vertically weighted soil water content measurements of two distributed in situ soil water content sensor networks. All three methods were successfully calibrated to determine field scale soil water content continuously at the ten sites. The hmf-method and the COSMIC-method had more similar calibration curves than the N0-method. The three methods performed similarly well in the validation and errors were within the uncertainty of neutron flux measurements despite observed differences in the calibration curves and variable model complexity. In addition, we found that the obtained calibration parameters NCOSMIC, N0 and NS showed a strong correlation with aboveground biomass.

  16. Altazimuth mount based dynamic calibration method for GNSS attitude measurement

    NASA Astrophysics Data System (ADS)

    Jiang, Nan; He, Tao; Sun, Shaohua; Gu, Qing

    2015-02-01

    As the key process to ensure the test accuracy and quality, the dynamic calibration of the GNSS attitude measuring instrument is often embarrassed by the lack of the rigid enough test platform and an accurate enough calibration reference. To solve the problems, a novel dynamic calibration method for GNSS attitude measurement based on altazimuth mount is put forward in this paper. The principle and implementation of this method are presented, and then the feasibility and usability of the method are analyzed in detail involving the applicability of the mount, calibrating precision, calibrating range, base line rigidity and the satellite signal involved factors. Furthermore, to verify and test the method, a confirmatory experiment is carried out with the survey ship GPS attitude measuring instrument, and the experimental results prove that it is a feasible way to the dynamic calibration for GNSS attitude measurement.

  17. Research radiometric calibration quantitative transfer methods between internal and external

    NASA Astrophysics Data System (ADS)

    Guo, Ju Guang; Ma, Yong hui; Zhang, Guang; Yang, Zhi hui

    2015-10-01

    This paper puts forward a method by realizing the internal and external radiation calibration transfer for infrared radiation characteristics quantitative measuring system. Through technological innovation and innovation application to establish a theoretical model of the corresponding radiated transfer method. This method can be well in engineering application for technology conversion process of radiometric calibration that with relatively simple and effective calibration in the half light path radiation instead of complex difficult whole optical path radiometric calibration. At the same time, it also will provide the basis of effective support to further carry out the target radiated characteristics quantitative measurement and application for ground type infrared radiated quantitative measuring system.

  18. 40 CFR 80.170 - Volumetric additive reconciliation (VAR), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2013 CFR

    2013-07-01

    ... ADDITIVES Detergent Gasoline § 80.170 Volumetric additive reconciliation (VAR), equipment calibration, and recordkeeping requirements. This section contains requirements for automated detergent blending facilities and hand-blending detergent facilities. All gasoline and all PRC intended for use in gasoline must...

  19. 40 CFR 80.157 - Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2013 CFR

    2013-07-01

    ... ADDITIVES Detergent Gasoline § 80.157 Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements. This section contains requirements for automated detergent blending facilities and hand-blending detergent facilities. All gasolines and all PRC intended for use in gasoline must...

  20. 40 CFR 80.157 - Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... ADDITIVES Detergent Gasoline § 80.157 Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements. This section contains requirements for automated detergent blending facilities and hand-blending detergent facilities. All gasolines and all PRC intended for use in gasoline must...

  1. 40 CFR 80.170 - Volumetric additive reconciliation (VAR), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... ADDITIVES Detergent Gasoline § 80.170 Volumetric additive reconciliation (VAR), equipment calibration, and recordkeeping requirements. This section contains requirements for automated detergent blending facilities and hand-blending detergent facilities. All gasoline and all PRC intended for use in gasoline must...

  2. Theoretical foundation, methods, and criteria for calibrating human vibration models using frequency response functions

    PubMed Central

    Dong, Ren G.; Welcome, Daniel E.; McDowell, Thomas W.; Wu, John Z.

    2015-01-01

    While simulations of the measured biodynamic responses of the whole human body or body segments to vibration are conventionally interpreted as summaries of biodynamic measurements, and the resulting models are considered quantitative, this study looked at these simulations from a different angle: model calibration. The specific aims of this study are to review and clarify the theoretical basis for model calibration, to help formulate the criteria for calibration validation, and to help appropriately select and apply calibration methods. In addition to established vibration theory, a novel theorem of mechanical vibration is also used to enhance the understanding of the mathematical and physical principles of the calibration. Based on this enhanced understanding, a set of criteria was proposed and used to systematically examine the calibration methods. Besides theoretical analyses, a numerical testing method is also used in the examination. This study identified the basic requirements for each calibration method to obtain a unique calibration solution. This study also confirmed that the solution becomes more robust if more than sufficient calibration references are provided. Practically, however, as more references are used, more inconsistencies can arise among the measured data for representing the biodynamic properties. To help account for the relative reliabilities of the references, a baseline weighting scheme is proposed. The analyses suggest that the best choice of calibration method depends on the modeling purpose, the model structure, and the availability and reliability of representative reference data. PMID:26740726

  3. Theoretical foundation, methods, and criteria for calibrating human vibration models using frequency response functions

    NASA Astrophysics Data System (ADS)

    Dong, Ren G.; Welcome, Daniel E.; McDowell, Thomas W.; Wu, John Z.

    2015-11-01

    While simulations of the measured biodynamic responses of the whole human body or body segments to vibration are conventionally interpreted as summaries of biodynamic measurements, and the resulting models are considered quantitative, this study looked at these simulations from a different angle: model calibration. The specific aims of this study are to review and clarify the theoretical basis for model calibration, to help formulate the criteria for calibration validation, and to help appropriately select and apply calibration methods. In addition to established vibration theory, a novel theorem of mechanical vibration is also used to enhance the understanding of the mathematical and physical principles of the calibration. Based on this enhanced understanding, a set of criteria was proposed and used to systematically examine the calibration methods. Besides theoretical analyses, a numerical testing method is also used in the examination. This study identified the basic requirements for each calibration method to obtain a unique calibration solution. This study also confirmed that the solution becomes more robust if more than sufficient calibration references are provided. Practically, however, as more references are used, more inconsistencies can arise among the measured data for representing the biodynamic properties. To help account for the relative reliabilities of the references, a baseline weighting scheme is proposed. The analyses suggest that the best choice of calibration method depends on the modeling purpose, the model structure, and the availability and reliability of representative reference data.

  4. Calibration methods for rotating shadowband irradiometers and optimizing the calibration duration

    NASA Astrophysics Data System (ADS)

    Jessen, Wilko; Wilbert, Stefan; Nouri, Bijan; Geuder, Norbert; Fritz, Holger

    2016-04-01

    Resource assessment for concentrated solar power (CSP) needs accurate direct normal irradiance (DNI) measurements. An option for such measurement campaigns is the use of thoroughly calibrated rotating shadowband irradiometers (RSIs). Calibration of RSIs and Si-sensors is complex because of the inhomogeneous spectral response of these sensors and incorporates the use of several correction functions. One calibration for a given atmospheric condition and air mass might not be suitable under different conditions. This paper covers procedures and requirements of two calibration methods for the calibration of rotating shadowband irradiometers. The necessary duration of acquisition of test measurements is examined with regard to the site-specific conditions at Plataforma Solar de Almería (PSA) in Spain. Seven data sets of long-term test measurements were collected. For each data set, calibration results of varying durations were compared to its respective long-term result. Our findings show that seasonal changes of environmental conditions are causing small but noticeable fluctuation of calibration results. Calibration results within certain periods (i.e. November to January and April to May) show a higher likelihood of deviation. These effects can partially be attenuated by including more measurements from outside these periods. Consequently, the duration of calibrations at PSA can now be selected depending on the time of year in which measurements commence.

  5. Simultaneous multi-headed imager geometry calibration method

    DOEpatents

    Tran, Vi-Hoa; Meikle, Steven Richard; Smith, Mark Frederick

    2008-02-19

    A method for calibrating multi-headed high sensitivity and high spatial resolution dynamic imaging systems, especially those useful in the acquisition of tomographic images of small animals. The method of the present invention comprises: simultaneously calibrating two or more detectors to the same coordinate system; and functionally correcting for unwanted detector movement due to gantry flexing.

  6. A robust method for determining calibration coefficients for VIIRS reflective solar bands

    NASA Astrophysics Data System (ADS)

    Ji, Qiang; McIntire, Jeffrey; Efremova, Boryana; Schwarting, Thomas; Oudrari, Hassan; Zeng, Jinan; Xiong, Xiaoxiong

    2015-09-01

    This paper presents a robust method for determining the calibration coefficients in polynomial calibration equations, and discusses the corresponding calibration uncertainties. An attenuator method that takes into account all measurements with and without an attenuator screen was used to restrict the impact of the absolute calibration of the light source. The originally proposed procedure attempts to simultaneously determine all unknowns nonlinearly using polynomial curve fitting. The newly proposed method divides the task into two simpler parts. For example, in the case of a quadratic calibration equation, the first part becomes a quadratic equation solely for the transmittance of attenuator, which has an analytical solution using three or four sets of measurements. Additionally, it is straightforward to determine the median value and the standard deviation of the transmittance from the solutions using all combinations of measured data points. In conjunction, the second part becomes a linear fit, with the ratio of the zeroth-order to first-order calibration coefficients as the intercept and the ratio of the second-order to first-order calibration coefficients as the slope. These ratios are unaffected by the absolute calibration of the light source and are then used in the calibration equation to calculate the first-order calibration coefficient. How the new method works is straightforward to visualize, which makes its results easier to verify. This is demonstrated using measurements from the Joint Polar Satellite System (JPSS) J1 Visible Infrared Imaging Radiometer Suite (VIIRS) reflective solar bands (RSB) pre-launch testing.

  7. Establishment of a primary reference solar cell calibration technique in Korea: methods, results and comparison with WPVS qualified laboratories

    NASA Astrophysics Data System (ADS)

    Ahn, SeungKyu; Ahn, SeJin; Yun, Jae Ho; Lee, Dong-Hoon; Winter, Stefan; Igari, Sanekazu; Yoon, KyungHoon

    2014-06-01

    A primary reference solar cell calibration technique recently established at the Korea Institute of Energy Research in Korea is introduced. This calibration technique is an indoor method that uses a highly collimated continuous-type solar simulator and absolute cavity radiometer traceable to the World Radiometric Reference. The results obtained using this calibration technique are shown with a precise uncertainty analysis, and the system configuration and calibration procedures are introduced. The calibration technique avoids overestimating the short-circuit current of a reference solar cell due to multiple reflections of incident simulator light using a novel method. In addition, the uncertainty analysis indicates that the calibration technique has an expanded uncertainty of approximately 0.7% with a coverage factor of k = 2 for a c-Si reference cell calibration. In addition, the developed primary reference solar cell calibration technique was compared with other techniques established in the World Photovoltaic Scale (WPVS) qualified calibration laboratories to verify its validity and reliability.

  8. Comparison of TLD calibration methods for 192Ir dosimetry.

    PubMed

    Haworth, Annette; Butler, Duncan J; Wilfert, Lisa; Ebert, Martin A; Todd, Stephen P; Hayton, Anna J M; Kron, Tomas

    2013-01-01

    For the purpose of dose measurement using a high-dose rate (192)Ir source, four methods of thermoluminescent dosimeter (TLD) calibration were investigated. Three of the four calibration methods used the (192)Ir source. Dwell times were calculated to deliver 1 Gy to the TLDs irradiated either in air or water. Dwell time calculations were confirmed by direct measurement using an ionization chamber. The fourth method of calibration used 6 MV photons from a medical linear accelerator, and an energy correction factor was applied to account for the difference in sensitivity of the TLDs in (192)Ir and 6 MV. The results of the four TLD calibration methods are presented in terms of the results of a brachytherapy audit where seven Australian centers irradiated three sets of TLDs in a water phantom. The results were in agreement within estimated uncertainties when the TLDs were calibrated with the (192)Ir source. Calibrating TLDs in a phantom similar to that used for the audit proved to be the most practical method and provided the greatest confidence in measured dose. When calibrated using 6 MV photons, the TLD results were consistently higher than the (192)Ir-calibrated TLDs, suggesting this method does not fully correct for the response of the TLDs when irradiated in the audit phantom. PMID:23318392

  9. Uncertainty Analysis of Inertial Model Attitude Sensor Calibration and Application with a Recommended New Calibration Method

    NASA Technical Reports Server (NTRS)

    Tripp, John S.; Tcheng, Ping

    1999-01-01

    Statistical tools, previously developed for nonlinear least-squares estimation of multivariate sensor calibration parameters and the associated calibration uncertainty analysis, have been applied to single- and multiple-axis inertial model attitude sensors used in wind tunnel testing to measure angle of attack and roll angle. The analysis provides confidence and prediction intervals of calibrated sensor measurement uncertainty as functions of applied input pitch and roll angles. A comparative performance study of various experimental designs for inertial sensor calibration is presented along with corroborating experimental data. The importance of replicated calibrations over extended time periods has been emphasized; replication provides independent estimates of calibration precision and bias uncertainties, statistical tests for calibration or modeling bias uncertainty, and statistical tests for sensor parameter drift over time. A set of recommendations for a new standardized model attitude sensor calibration method and usage procedures is included. The statistical information provided by these procedures is necessary for the uncertainty analysis of aerospace test results now required by users of industrial wind tunnel test facilities.

  10. Calibration method for video and radiation imagers

    DOEpatents

    Cunningham, Mark F.; Fabris, Lorenzo; Gee, Timothy F.; Goddard, Jr., James S.; Karnowski, Thomas P.; Ziock, Klaus-peter

    2011-07-05

    The relationship between the high energy radiation imager pixel (HERIP) coordinate and real-world x-coordinate is determined by a least square fit between the HERIP x-coordinate and the measured real-world x-coordinates of calibration markers that emit high energy radiation imager and reflect visible light. Upon calibration, a high energy radiation imager pixel position may be determined based on a real-world coordinate of a moving vehicle. Further, a scale parameter for said high energy radiation imager may be determined based on the real-world coordinate. The scale parameter depends on the y-coordinate of the moving vehicle as provided by a visible light camera. The high energy radiation imager may be employed to detect radiation from moving vehicles in multiple lanes, which correspondingly have different distances to the high energy radiation imager.

  11. A Bionic Polarization Navigation Sensor and Its Calibration Method.

    PubMed

    Zhao, Huijie; Xu, Wujian

    2016-01-01

    The polarization patterns of skylight which arise due to the scattering of sunlight in the atmosphere can be used by many insects for deriving compass information. Inspired by insects' polarized light compass, scientists have developed a new kind of navigation method. One of the key techniques in this method is the polarimetric sensor which is used to acquire direction information from skylight. In this paper, a polarization navigation sensor is proposed which imitates the working principles of the polarization vision systems of insects. We introduce the optical design and mathematical model of the sensor. In addition, a calibration method based on variable substitution and non-linear curve fitting is proposed. The results obtained from the outdoor experiments provide support for the feasibility and precision of the sensor. The sensor's signal processing can be well described using our mathematical model. A relatively high degree of accuracy in polarization measurement can be obtained without any error compensation. PMID:27527171

  12. System and method for calibrating a rotary absolute position sensor

    NASA Technical Reports Server (NTRS)

    Davis, Donald R. (Inventor); Permenter, Frank Noble (Inventor); Radford, Nicolaus A (Inventor)

    2012-01-01

    A system includes a rotary device, a rotary absolute position (RAP) sensor generating encoded pairs of voltage signals describing positional data of the rotary device, a host machine, and an algorithm. The algorithm calculates calibration parameters usable to determine an absolute position of the rotary device using the encoded pairs, and is adapted for linearly-mapping an ellipse defined by the encoded pairs to thereby calculate the calibration parameters. A method of calibrating the RAP sensor includes measuring the rotary position as encoded pairs of voltage signals, linearly-mapping an ellipse defined by the encoded pairs to thereby calculate the calibration parameters, and calculating an absolute position of the rotary device using the calibration parameters. The calibration parameters include a positive definite matrix (A) and a center point (q) of the ellipse. The voltage signals may include an encoded sine and cosine of a rotary angle of the rotary device.

  13. Method and apparatus for calibrating a particle emissions monitor

    DOEpatents

    Flower, William L.; Renzi, Ronald F.

    1998-07-07

    The instant invention discloses method and apparatus for calibrating particulate emissions monitors, in particular, and sampling probes, in general, without removing the instrument from the system being monitored. A source of one or more specific metals in aerosol (either solid or liquid) or vapor form is housed in the instrument. The calibration operation is initiated by moving a focusing lens, used to focus a light beam onto an analysis location and collect the output light response, from an operating position to a calibration position such that the focal point of the focusing lens is now within a calibration stream issuing from a calibration source. The output light response from the calibration stream can be compared to that derived from an analysis location in the operating position to more accurately monitor emissions within the emissions flow stream.

  14. Method and apparatus for calibrating a particle emissions monitor

    DOEpatents

    Flower, W.L.; Renzi, R.F.

    1998-07-07

    The invention discloses a method and apparatus for calibrating particulate emissions monitors, in particular, sampling probes, and in general, without removing the instrument from the system being monitored. A source of one or more specific metals in aerosol (either solid or liquid) or vapor form is housed in the instrument. The calibration operation is initiated by moving a focusing lens, used to focus a light beam onto an analysis location and collect the output light response, from an operating position to a calibration position such that the focal point of the focusing lens is now within a calibration stream issuing from a calibration source. The output light response from the calibration stream can be compared to that derived from an analysis location in the operating position to more accurately monitor emissions within the emissions flow stream. 6 figs.

  15. A novel calibration method of CCD camera for LAMOST

    NASA Astrophysics Data System (ADS)

    Gu, Yonggang; Jin, Yi; Zhai, Chao

    2012-09-01

    Large Sky Area Multi-object Fiber Spectroscopic Telescope - LAMOST, with a 1.75m-diameter focal plane on which 4000 optical fibers are arranged, is one of major scientific projects in China. During the surveying process of LAMOST, the optical imaging system makes the astrometric objects be imaged in the focal plane, and the optical fiber positioning system controls the 4000 fibers to be aligned with these objects and obtain their spectrum. In order to correct the positioning error of these optical fibers, the CCD camera is used to detect these fibers’ position in the way of close-range photogrammetry. As we all know, the calibration quality of the CCD camera is one of the most important factors for detection precision. However, the camera calibration has two following problems in the field work of LAMOST. First, the camera parameters are not stable due to the changes of on-site work environment and the vibration during movement. So, the CCD camera must be on-line calibrated. Second, a large-size high-precision calibration target is needed to calibrate the camera, for the focal plane is very big. Making such a calibration target, it is very difficult and costly. Meanwhile, the large calibration target is hard to be fixed on LAMOST because of the space constraint. In this paper, an improved bundle adjustment self-calibration method is proposed to solve the two problems above. The results of experiment indicate that this novel calibration method needs only a few control points while the traditional calibration methods need much more control points to get the same accuracy. So the method could realize the on-line high-precision calibration of CCD camera for LAMOST.

  16. Method of Noncontact Calibration of the Robotic Ultrasonic Tomograph

    NASA Astrophysics Data System (ADS)

    Borikov, V. N.; Galtseva, O. V.; Filippov, G. A.

    2016-01-01

    The method of calibration of robotic ultrasonic tomograph with the construction of the trajectory of movement of the robot-manipulator on the object of control by using 3D- scanner is described. This method can significantly accelerate the process of calibration of tomograph and prevent possible displacement of the object during calibration. The algorithm of transition from use of a contact method of calibration of the tomograph to noncontact calibration is offered. Experimental data of application of this algorithm show a positive result: the time of research of object considerably decreases. Results of researches prove the practical relevance of the presented work and high efficiency of application of robotic ultrasonic tomography for nondestructive testing of objects of different forms.

  17. 40 CFR 80.157 - Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2011 CFR

    2011-07-01

    ... limited number of specified people those who have the ability to alter or delete the records. In addition... package has the same viscosity as the previous detergent package. Detergent package change calibrations.... Detergent so used must be accurately and separately measured, either through the use of a separate...

  18. 40 CFR 80.157 - Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2010 CFR

    2010-07-01

    ... limited number of specified people those who have the ability to alter or delete the records. In addition... package has the same viscosity as the previous detergent package. Detergent package change calibrations.... Detergent so used must be accurately and separately measured, either through the use of a separate...

  19. New method for calibration of a computer controlled CRT: feedback calibration

    NASA Astrophysics Data System (ADS)

    Wang, Jingping; Zhu, Zhengfang; Zhao, Dazun

    1994-08-01

    There are many formulas to describe the input and output relationship of CRT. As the radiant output of CRT submits to the statistical distribution, the prediction of the colorimetry from these formulas is not good. Some deviation of the brightness are as high as 50%. There is a need to find a new method to calibrate CRT. We have set up an automatic measurement and calibration system with PHOTO 1980B, IEEE 4888 interface board, CRT and a computer.

  20. Vanishing feature constraints calibration method for binocular vision sensor.

    PubMed

    Wei, Zhenzhong; Liu, Xiaokun

    2015-07-27

    Based on analyzing the measurement model of binocular vision sensor, we proposed a new flexible calibration method for binocular vision sensor using a planar target with several parallel lines. It only requires the sensor to observe the planar target at a few (at least two) different orientations. Relying on vanishing feature constraints and spacing constraints of parallel lines, linear method and nonlinear optimization are combined to estimate the structure parameters of binocular vision sensor. Linear method achieves the separation of the rotation matrix and translation vector which reduces the complexity of computation; Nonlinear algorithm ensures the calibration results for the global optimization. Towards the factors that affect the accuracy of the calibration, theoretical analysis and computer simulation are carried out respectively consequence in qualitative analysis and quantitative result. Real data shows that the accuracy of the proposed calibration method is about 0.040mm with the working distance of 800mm and the view field of 300 × 300mm. The comparison with Bougust toolbox and the method based on known length indicates that the proposed calibration method is precise and is efficient and convenient as its simple calculation and easy operation, especially for onsite calibration and self-calibration. PMID:26367553

  1. Automatable on-line generation of calibration curves and standard additions in solution-cathode glow discharge optical emission spectrometry

    NASA Astrophysics Data System (ADS)

    Schwartz, Andrew J.; Ray, Steven J.; Hieftje, Gary M.

    2015-03-01

    Two methods are described that enable on-line generation of calibration standards and standard additions in solution-cathode glow discharge optical emission spectrometry (SCGD-OES). The first method employs a gradient high-performance liquid chromatography pump to perform on-line mixing and delivery of a stock standard, sample solution, and diluent to achieve a desired solution composition. The second method makes use of a simpler system of three peristaltic pumps to perform the same function of on-line solution mixing. Both methods can be computer-controlled and automated, and thereby enable both simple and standard-addition calibrations to be rapidly performed on-line. Performance of the on-line approaches is shown to be comparable to that of traditional methods of sample preparation, in terms of calibration curves, signal stability, accuracy, and limits of detection. Potential drawbacks to the on-line procedures include signal lag between changes in solution composition and pump-induced multiplicative noise. Though the new on-line methods were applied here to SCGD-OES to improve sample throughput, they are not limited in application to only SCGD-OES-any instrument that samples from flowing solution streams (flame atomic absorption spectrometry, ICP-OES, ICP-mass spectrometry, etc.) could benefit from them.

  2. Scatterometer-Calibrated Stability Verification Method

    NASA Technical Reports Server (NTRS)

    McWatters, Dalia A.; Cheetham, Craig M.; Huang, Shouhua; Fischman, Mark A.; CHu, Anhua J.; Freedman, Adam P.

    2011-01-01

    The requirement for scatterometer-combined transmit-receive gain variation knowledge is typically addressed by sampling a portion of the transmit signal, attenuating it with a known-stable attenuation, and coupling it into the receiver chain. This way, the gain variations of the transmit and receive chains are represented by this loop-back calibration signal, and can be subtracted from the received remote radar echo. Certain challenges are presented by this process, such as transmit and receive components that are outside of this loop-back path and are not included in this calibration, as well as the impracticality for measuring the transmit and receive chains stability and post fabrication separately, without the resulting measurement errors from the test set up exceeding the requirement for the flight instrument. To cover the RF stability design challenge, the portions of the scatterometer that are not calibrated by the loop-back, (e.g., attenuators, switches, diplexers, couplers, and coaxial cables) are tightly thermally controlled, and have been characterized over temperature to contribute less than 0.05 dB of calibration error over worst-case thermal variation. To address the verification challenge, including the components that are not calibrated by the loop-back, a stable fiber optic delay line (FODL) was used to delay the transmitted pulse, and to route it into the receiver. In this way, the internal loopback signal amplitude variations can be compared to the full transmit/receive external path, while the flight hardware is in the worst-case thermal environment. The practical delay for implementing the FODL is 100 s. The scatterometer pulse width is 1 ms so a test mode was incorporated early in the design phase to scale the 1 ms pulse at 100-Hz pulse repetition interval (PRI), by a factor of 18, to be a 55 s pulse with 556 s PRI. This scaling maintains the duty cycle, thus maintaining a representative thermal state for the RF components. The FODL consists

  3. Evaluation of two gas-dilution methods for instrument calibration

    NASA Technical Reports Server (NTRS)

    Evans, A., Jr.

    1977-01-01

    Two gas dilution methods were evaluated for use in the calibration of analytical instruments used in air pollution studies. A dual isotope fluorescence carbon monoxide analyzer was used as the transfer standard. The methods are not new but some modifications are described. The rotary injection gas dilution method was found to be more accurate than the closed loop method. Results by the two methods differed by 5 percent. This could not be accounted for by the random errors in the measurements. The methods avoid the problems associated with pressurized cylinders. Both methods have merit and have found a place in instrument calibration work.

  4. Bond additivity corrections for quantum chemistry methods

    SciTech Connect

    C. F. Melius; M. D. Allendorf

    1999-04-01

    In the 1980's, the authors developed a bond-additivity correction procedure for quantum chemical calculations called BAC-MP4, which has proven reliable in calculating the thermochemical properties of molecular species, including radicals as well as stable closed-shell species. New Bond Additivity Correction (BAC) methods have been developed for the G2 method, BAC-G2, as well as for a hybrid DFT/MP2 method, BAC-Hybrid. These BAC methods use a new form of BAC corrections, involving atomic, molecular, and bond-wise additive terms. These terms enable one to treat positive and negative ions as well as neutrals. The BAC-G2 method reduces errors in the G2 method due to nearest-neighbor bonds. The parameters within the BAC-G2 method only depend on atom types. Thus the BAC-G2 method can be used to determine the parameters needed by BAC methods involving lower levels of theory, such as BAC-Hybrid and BAC-MP4. The BAC-Hybrid method should scale well for large molecules. The BAC-Hybrid method uses the differences between the DFT and MP2 as an indicator of the method's accuracy, while the BAC-G2 method uses its internal methods (G1 and G2MP2) to provide an indicator of its accuracy. Indications of the average error as well as worst cases are provided for each of the BAC methods.

  5. A self-sufficient method for calibration of Varian electronic portal imaging device

    NASA Astrophysics Data System (ADS)

    Sun, Baozhou; Yaddanapudi, Sridhar; Goddu, Sreekrishna M.; Mutic, Sasa

    2015-01-01

    Electronic portal imaging device (EPID) is currently used for dosimetric verification of IMRT fields and linac quality assurance (QA). It is critical to understand the dosimetric response and perform an accurate and robust calibration of EPID. We present the implementation of an efficient method for the calibration and the validation of a Varian EPID, which relies only on data collected with that specific device. The calibration method is based on images obtained with five shifts of EPID panel. With this method, the relative gain (sensitivity) of each element of a detector matrix is calculated and applied on top of the calibration determined with the flood-field procedure. The calibration procedure was verified using a physical wedge inserted in the beam line and the corrected profile shows consistent results with the measurements using a calibrated 2D array. This method does not rely on the beam profile used in the flood-field calibration process, which allows EPID calibration in 10 minutes with no additional equipment compared to at least 2 hours to obtain beam profile and scanning beam equipment requirement with the conventional method.

  6. Interferometric SAR phase difference calibration: Methods and results

    SciTech Connect

    Bickel, D.L.; Hensley, W.H.

    1993-12-31

    This paper addresses the steps necessary to determine and maintain the phase calibration of a two-channel interferometric synthetic aperture radar (IFSAR). The method, setup, and accuracy of four different calibration techniques are compared. The most novel technique involves pointing the interferometric baseline at nadir and imaging a lake surface. The other techniques include measuring various flat surfaces in traditional side-looking IFSAR maps, in-flight closed-loop calibration path measurements, and static laboratory measurements. Initial results indicate that, using combinations of these measurements, it is possible to maintain the interferometric phase calibration of Sandia National Laboratories` K{sub U} Band IFSAR to better than 3 degrees. The time variability of various parts of the calibration and requirements for recalibration are also discussed.

  7. Method calibration of the model 13145 infrared target projectors

    NASA Astrophysics Data System (ADS)

    Huang, Jianxia; Gao, Yuan; Han, Ying

    2014-11-01

    The SBIR Model 13145 Infrared Target Projectors ( The following abbreviation Evaluation Unit ) used for characterizing the performances of infrared imaging system. Test items: SiTF, MTF, NETD, MRTD, MDTD, NPS. Infrared target projectors includes two area blackbodies, a 12 position target wheel, all reflective collimator. It provide high spatial frequency differential targets, Precision differential targets imaged by infrared imaging system. And by photoelectricity convert on simulate signal or digital signal. Applications software (IR Windows TM 2001) evaluate characterizing the performances of infrared imaging system. With regards to as a whole calibration, first differently calibration for distributed component , According to calibration specification for area blackbody to calibration area blackbody, by means of to amend error factor to calibration of all reflective collimator, radiance calibration of an infrared target projectors using the SR5000 spectral radiometer, and to analyze systematic error. With regards to as parameter of infrared imaging system, need to integrate evaluation method. According to regulation with -GJB2340-1995 General specification for military thermal imaging sets -testing parameters of infrared imaging system, the results compare with results from Optical Calibration Testing Laboratory . As a goal to real calibration performances of the Evaluation Unit.

  8. Method for Accurately Calibrating a Spectrometer Using Broadband Light

    NASA Technical Reports Server (NTRS)

    Simmons, Stephen; Youngquist, Robert

    2011-01-01

    A novel method has been developed for performing very fine calibration of a spectrometer. This process is particularly useful for modern miniature charge-coupled device (CCD) spectrometers where a typical factory wavelength calibration has been performed and a finer, more accurate calibration is desired. Typically, the factory calibration is done with a spectral line source that generates light at known wavelengths, allowing specific pixels in the CCD array to be assigned wavelength values. This method is good to about 1 nm across the spectrometer s wavelength range. This new method appears to be accurate to about 0.1 nm, a factor of ten improvement. White light is passed through an unbalanced Michelson interferometer, producing an optical signal with significant spectral variation. A simple theory can be developed to describe this spectral pattern, so by comparing the actual spectrometer output against this predicted pattern, errors in the wavelength assignment made by the spectrometer can be determined.

  9. Self-calibration of terrestrial laser scanners: selection of the best geometric additional parameters

    NASA Astrophysics Data System (ADS)

    Lerma, J. L.; García-San-Miguel, D.

    2014-05-01

    Systematic errors are present in laser scanning system observations due to manufacturer imperfections, wearing over time, vibrations, changing environmental conditions and, last but not least, involuntary hits. To achieve maximum quality and rigorous measurements from terrestrial laser scanners, a least squares estimation of additional calibration parameters can be used to model the a priori unknown systematic errors and therefore improve output observations. The selection of the right set of additional parameters is not trivial and requires laborious statistical analysis. Based on this requirement, this article presents an approach to determine the best set of additional parameters which provides the best mathematical solution based on a dimensionless quality index. The best set of additional parameters is the one which provides the maximum quality index (i.e. minimum value) for the group of observables, exterior orientation parameters and reference points. Calibration performance is tested using both a phase shift continuous wave scanner, FARO PHOTON 880, and a pulse-based time-of-flight system, Leica HDS3000. The improvement achieved after the geometric calibration is 30% for the former and 70% for the latter.

  10. Bond additivity corrections for quantum chemistry methods

    SciTech Connect

    Melius, C.F.; Allendorf, M.D.

    2000-03-23

    New bond additivity correction (BAC) methods have been developed for the G2 method, BAC-G2, as well as for a hybrid density functional theory (DFT) Moller-Plesset (MP)2 method, BAC-hybrid. These BAC methods use a new form of BAC corrections, involving atomic, molecular, and bond-wise additive terms. These terms enable one to treat positive and negative ions as well as neutrals. The BAC-G2 method reduces errors in the G2 method due to nearest-neighbor bonds. The parameters within the BAC-G2 method only depend on atom types. Thus the BAC-G2 method can be used to determine the parameters needed by BAC methods involving lower levels of theory, such as BAC-hybrid and BAC-MP4. The BAC-hybrid method is expected to scale well for large molecules. The BAC-hybrid method uses the differences between the DFT and MP2 predictions as an indication of the method's accuracy, whereas the BAC-G2 method uses its internal methods (G1 and G2MP2) to accomplish this. A statistical analysis of the error in each of the methods is presented on the basis of calculations performed for large sets (more than 120) of molecules.

  11. Two laboratory methods for the calibration of GPS speed meters

    NASA Astrophysics Data System (ADS)

    Bai, Yin; Sun, Qiao; Du, Lei; Yu, Mei; Bai, Jie

    2015-01-01

    The set-ups of two calibration systems are presented to investigate calibration methods of GPS speed meters. The GPS speed meter calibrated is a special type of high accuracy speed meter for vehicles which uses Doppler demodulation of GPS signals to calculate the measured speed of a moving target. Three experiments are performed: including simulated calibration, field-test signal replay calibration, and in-field test comparison with an optical speed meter. The experiments are conducted at specific speeds in the range of 40-180 km h-1 with the same GPS speed meter as the device under calibration. The evaluation of measurement results validates both methods for calibrating GPS speed meters. The relative deviations between the measurement results of the GPS-based high accuracy speed meter and those of the optical speed meter are analyzed, and the equivalent uncertainty of the comparison is evaluated. The comparison results justify the utilization of GPS speed meters as reference equipment if no fewer than seven satellites are available. This study contributes to the widespread use of GPS-based high accuracy speed meters as legal reference equipment in traffic speed metrology.

  12. A New Online Calibration Method for Multidimensional Computerized Adaptive Testing.

    PubMed

    Chen, Ping; Wang, Chun

    2016-09-01

    Multidimensional-Method A (M-Method A) has been proposed as an efficient and effective online calibration method for multidimensional computerized adaptive testing (MCAT) (Chen & Xin, Paper presented at the 78th Meeting of the Psychometric Society, Arnhem, The Netherlands, 2013). However, a key assumption of M-Method A is that it treats person parameter estimates as their true values, thus this method might yield erroneous item calibration when person parameter estimates contain non-ignorable measurement errors. To improve the performance of M-Method A, this paper proposes a new MCAT online calibration method, namely, the full functional MLE-M-Method A (FFMLE-M-Method A). This new method combines the full functional MLE (Jones & Jin in Psychometrika 59:59-75, 1994; Stefanski & Carroll in Annals of Statistics 13:1335-1351, 1985) with the original M-Method A in an effort to correct for the estimation error of ability vector that might otherwise adversely affect the precision of item calibration. Two correction schemes are also proposed when implementing the new method. A simulation study was conducted to show that the new method generated more accurate item parameter estimation than the original M-Method A in almost all conditions. PMID:26608960

  13. Method to calibrate fission chambers in Campbelling mode

    SciTech Connect

    Benoit Geslot; Troy C. Unruh; Philippe Filliatre; Christian Jammes; Jacques Di Salvo; Stéphane Bréaud; Jean-François Villard

    2011-06-01

    Fission chambers are neutron detectors which are widely used to instrument experimental reactors such as material testing reactors or zero power reactors. In the presence of a high level mixed gamma and neutron flux, fission chambers can be operated in Campbelling mode (also known as 'fluctuation mode' or 'mean square voltage mode') to provide reliable and precise neutron related measurements. Fission chamber calibration in Campbelling mode (in terms of neutron flux) is usually done empirically using a calibrated reference detector. A major drawback of this method is that calibration measurements have to be performed in a neutron environment very similar to the one in which the calibrated detector will be used afterwards. What we propose here is a different approach based on characterizing the fission chamber response in terms of fission rate. This way, the detector calibration coefficient is independent from the neutron spectrum and can be determined prior to the experiment. The fissile deposit response to the neutron spectrum can then be assessed independently by other means (experimental or numerical). In this paper, the response of CEA made miniature fission chambers in Campbelling mode is studied. We use a theoretical model of the signal to calculate the calibration coefficient. Input parameters of the model come from statistical distribution of individual pulses. Supporting measurements have been made in the CEA Cadarache zero power reactor MINERVE. Results are compared to an empirical Campbelling mode calibration.

  14. Method of calibrating an interferometer and reducing its systematic noise

    NASA Technical Reports Server (NTRS)

    Hammer, Philip D. (Inventor)

    1997-01-01

    Methods of operation and data analysis for an interferometer so as to eliminate the errors contributed by non-responsive or unstable pixels, interpixel gain variations that drift over time, and spurious noise that would otherwise degrade the operation of the interferometer are disclosed. The methods provide for either online or post-processing calibration. The methods apply prescribed reversible transformations that exploit the physical properties of interferograms obtained from said interferometer to derive a calibration reference signal for subsequent treatment of said interferograms for interpixel gain variations. A self-consistent approach for treating bad pixels is incorporated into the methods.

  15. Site characterization for calibration of radiometric sensors using vicarious method

    NASA Astrophysics Data System (ADS)

    Parihar, Shailesh; Rathore, L. S.; Mohapatra, M.; Sharma, A. K.; Mitra, A. K.; Bhatla, R.; Singh, R. S.; Desai, Yogdeep; Srivastava, Shailendra S.

    2016-05-01

    Radiometric performances of earth observation satellite/sensors vary from ground pre-launch calibration campaign to post launch period extended to lifetime of the satellite due to launching vibrations. Therefore calibration is carried out worldwide through various methods throughout satellite lifetime. In India Indian Space Research Organization (ISRO) calibrates the sensor of Resourcesat-2 satellite by vicarious method. One of these vicarious calibration methods is the reflectance-based approach that is applied in this study for radiometric calibration of sensors on-board Resouresat-2 satellite. The results of ground-based measurement of atmospheric conditions and surface reflectance are made at Bap, Rajasthan Calibration/Validation (Cal/Val) site. Cal/Val observations at site were carried out with hyper-spectral Spectroradiometer covering spectral range of 350nm- 2500nm for radiometric characterization of the site. The Sunphotometer/Ozonometer for measuring the atmospheric parameters has also been used. The calibrated radiance is converted to absolute at-sensor spectral reflectance and Top-Of-Atmosphere (TOA) radiance. TOA radiance was computed using radiative transfer model `Second simulation of the satellite signal in the solar spectrum' (6S), which can accurately simulate the problems introduced by the presence of the atmosphere along the path from Sun to target (surface) to Sensor. The methodology for band averaged reflectance retrieval and spectral reflectance fitting process are described. Then the spectral reflectance and atmospheric parameters are put into 6S code to predict TOA radiance which compare with Resourcesat-2 radiance. Spectral signature and its reflectance ratio indicate the uniformity of the site. Thus the study proves that the selected site is suitable for vicarious calibration of sensor of Resourcesat-2. Further the study demonstrates the procedure for similar exercise for site selection for Cal/Val analysis of other satellite over India

  16. Method and apparatus for calibrating a linear variable differential transformer

    DOEpatents

    Pokrywka, Robert J.

    2005-01-18

    A calibration apparatus for calibrating a linear variable differential transformer (LVDT) having an armature positioned in au LVDT armature orifice, and the armature able to move along an axis of movement. The calibration apparatus includes a heating mechanism with an internal chamber, a temperature measuring mechanism for measuring the temperature of the LVDT, a fixture mechanism with an internal chamber for at least partially accepting the LVDT and for securing the LVDT within the heating mechanism internal chamber, a moving mechanism for moving the armature, a position measurement mechanism for measuring the position of the armature, and an output voltage measurement mechanism. A method for calibrating an LVDT, including the steps of: powering the LVDT; heating the LVDT to a desired temperature; measuring the position of the armature with respect to the armature orifice; and measuring the output voltage of the LVDT.

  17. Method and Appartus for Calibrating a Linear Variable Differential Transformer

    SciTech Connect

    Pokrywka, Robert J.

    2005-01-18

    A calibration apparatus for calibrating a linear variable differential transformer (LVDT) having an armature positioned in au LVDT armature orifice, and the armature able to move along an axis of movement. The calibration apparatus includes a heating mechanism with an internal chamber, a temperature measuring mechanism for measuring the temperature of the LVDT, a fixture mechanism with an internal chamber for at least partially accepting the LVDT and for securing the LVDT within the heating mechanism internal chamber, a moving mechanism for moving the armature, a position measurement mechanism for measuring the position of the armature, and an output voltage measurement mechanism. A method for calibrating an LVDT, including the steps of powering the LVDT; heating the LVDT to a desired temperature; measuring the position of the armature with respect to the armature orifice; and measuring the output voltage of the LVDT.

  18. An Accurate Projector Calibration Method Based on Polynomial Distortion Representation

    PubMed Central

    Liu, Miao; Sun, Changku; Huang, Shujun; Zhang, Zonghua

    2015-01-01

    In structure light measurement systems or 3D printing systems, the errors caused by optical distortion of a digital projector always affect the precision performance and cannot be ignored. Existing methods to calibrate the projection distortion rely on calibration plate and photogrammetry, so the calibration performance is largely affected by the quality of the plate and the imaging system. This paper proposes a new projector calibration approach that makes use of photodiodes to directly detect the light emitted from a digital projector. By analyzing the output sequence of the photoelectric module, the pixel coordinates can be accurately obtained by the curve fitting method. A polynomial distortion representation is employed to reduce the residuals of the traditional distortion representation model. Experimental results and performance evaluation show that the proposed calibration method is able to avoid most of the disadvantages in traditional methods and achieves a higher accuracy. This proposed method is also practically applicable to evaluate the geometric optical performance of other optical projection system. PMID:26492247

  19. A new full pose measurement method for robot calibration.

    PubMed

    Nguyen, Hoai-Nhan; Zhou, Jian; Kang, Hee-Jun

    2013-01-01

    Identification of robot kinematic errors during the calibration process often requires accurate full pose measurements (position and orientation) of robot end-effectors in Cartesian space. This paper proposes a new method of full pose measurement of robot end-effectors for calibration. This method is based on an analysis of the features of a set of target points (placed on a rotating end-effector) on a circular trajectory. The accurate measurement is validated by computational simulation results from the Puma robot. Moreover, experimental calibration and validation results for the Hyundai HA-06 robot prove the effectiveness, correctness, and reliability of the proposed method. This method can be applied to robots that have entirely revolute joints or to robots for which only the last joint is revolute. PMID:23863856

  20. A Comparative Study of IRT Fixed Parameter Calibration Methods

    ERIC Educational Resources Information Center

    Kim, Seonghoon

    2006-01-01

    This article provides technical descriptions of five fixed parameter calibration (FPC) methods, which were based on marginal maximum likelihood estimation via the EM algorithm, and evaluates them through simulation. The five FPC methods described are distinguished from each other by how many times they update the prior ability distribution and by…

  1. An improved method for determining force balance calibration accuracy

    NASA Astrophysics Data System (ADS)

    Ferris, Alice T.

    The results of an improved statistical method used at Langley Research Center for determining and stating the accuracy of a force balance calibration are presented. The application of the method for initial loads, initial load determination, auxiliary loads, primary loads, and proof loads is described. The data analysis is briefly addressed.

  2. Comparison of methods for calibrating AVIRIS data to ground reflectance

    NASA Technical Reports Server (NTRS)

    Clark, Roger N.; Swayze, Gregg; Heidebrecht, Kathy; Goetz, Alexander F. H.; Green, Robert O.

    1993-01-01

    We are comparing three basic methods of calibrating AVIRIS data to ground reflectance: (1) atmospheric radiative transfer models with the solar flux can be used to calibrate AVIRIS radiance data (Specific methods include the University of Colorado CSES ARP and ATREM algorithms); (2) Robert Green's modified MODTRAN and AVIRIS radiance model (This method is similar to 1 but differs in that the solar radiance is bypassed, so any errors in the solar flux are canceled, too); and (3) ground calibration using known sites in the AVIRIS scene. We are using 1992AVIRIS data over Cuprite, Nevada, and Blackhawk Island, Wisconsin, as our test scenes. Both these sites have extensive field measurements. The Cuprite site had a very clear atmosphere, thus path radiance was dominated by Rayleigh scattering with little or no flux beyond 1 micron. The Blackhawk site has more aerosols, with significant path radiance flux beyond 2 micron.

  3. Universal calculation formula and calibration method in Fourier transform profilometry

    SciTech Connect

    Wen Yongfu; Li Sikun; Cheng Haobo; Su Xianyu; Zhang Qican

    2010-12-01

    We propose a universal calculation formula of Fourier transform profilometry and give a strict theoretical analysis about the phase-height mapping relation. As the request on the experimental setup of the universal calculation formula is unconfined, the projector and the camera can be located arbitrarily to get better fringe information, which makes the operation flexible. The phase-height calibration method under the universal condition is proposed, which can avoid measuring the system parameters directly. It makes the system easy to manipulate and improves the measurement velocity. A computer simulation and experiment are conducted to verify its validity. The calculation formula and calibration method have been applied to measure an object of 22.00 mm maximal height. The relative error of the measurement result is only 0.59%. The experimental results prove that the three-dimensional shape of tested objects can be reconstructed exactly by using the calculation formula and calibration method, and the system has better universality.

  4. Accurate projector calibration method by using an optical coaxial camera.

    PubMed

    Huang, Shujun; Xie, Lili; Wang, Zhangying; Zhang, Zonghua; Gao, Feng; Jiang, Xiangqian

    2015-02-01

    Digital light processing (DLP) projectors have been widely utilized to project digital structured-light patterns in 3D imaging systems. In order to obtain accurate 3D shape data, it is important to calibrate DLP projectors to obtain the internal parameters. The existing projector calibration methods have complicated procedures or low accuracy of the obtained parameters. This paper presents a novel method to accurately calibrate a DLP projector by using an optical coaxial camera. The optical coaxial geometry is realized by a plate beam splitter, so the DLP projector can be treated as a true inverse camera. A plate having discrete markers on the surface is used to calibrate the projector. The corresponding projector pixel coordinate of each marker on the plate is determined by projecting vertical and horizontal sinusoidal fringe patterns on the plate surface and calculating the absolute phase. The internal parameters of the DLP projector are obtained by the corresponding point pair between the projector pixel coordinate and the world coordinate of discrete markers. Experimental results show that the proposed method can accurately calibrate the internal parameters of a DLP projector. PMID:25967789

  5. Effusion plate using additive manufacturing methods

    DOEpatents

    Johnson, Thomas Edward; Keener, Christopher Paul; Ostebee, Heath Michael; Wegerif, Daniel Gerritt

    2016-04-12

    Additive manufacturing techniques may be utilized to construct effusion plates. Such additive manufacturing techniques may include defining a configuration for an effusion plate having one or more internal cooling channels. The manufacturing techniques may further include depositing a powder into a chamber, applying an energy source to the deposited powder, and consolidating the powder into a cross-sectional shape corresponding to the defined configuration. Such methods may be implemented to construct an effusion plate having one or more channels with a curved cross-sectional geometry.

  6. An Automated Method for Ozonesonde Calibration: New Insights

    NASA Technical Reports Server (NTRS)

    Schmidlin, F. J.; Hoegger, Bruno A.; Levrat, Gilbert; Baldwin, Tony

    2008-01-01

    An automated method for preparation of the electrochemical concentration cell (ECC) ozonesonde is presented. Development of a computer-controlled system for preparation and calibration of the ECC is an improvement over the manual preparation method, and reduces subjectivity considerably. Preparation measurements in digital form aids analysis of the ECC before release and enhances post-flight data certification. Calibration of ozonesondes over a range of ozone concentrations between 0 mPA and 30 mPA is discussed. This presentation describes the automatic system, gives examples of calibrations. The automated system enables comparison of varying potassium iodide (KI) concentrations that should allow adjustment of earlier ozonesonde data obtained with different KT concentrations used since 1970, i.e., 2, 1.5, 1, and 0.5 percent. Preliminary results indicate ECC accuracy has a strong dependence on the electrolyte concentration and should not be considered linear with altitude.

  7. An Expectation-Maximization Method for Calibrating Synchronous Machine Models

    SciTech Connect

    Meng, Da; Zhou, Ning; Lu, Shuai; Lin, Guang

    2013-07-21

    The accuracy of a power system dynamic model is essential to its secure and efficient operation. Lower confidence in model accuracy usually leads to conservative operation and lowers asset usage. To improve model accuracy, this paper proposes an expectation-maximization (EM) method to calibrate the synchronous machine model using phasor measurement unit (PMU) data. First, an extended Kalman filter (EKF) is applied to estimate the dynamic states using measurement data. Then, the parameters are calculated based on the estimated states using maximum likelihood estimation (MLE) method. The EM method iterates over the preceding two steps to improve estimation accuracy. The proposed EM method’s performance is evaluated using a single-machine infinite bus system and compared with a method where both state and parameters are estimated using an EKF method. Sensitivity studies of the parameter calibration using EM method are also presented to show the robustness of the proposed method for different levels of measurement noise and initial parameter uncertainty.

  8. A stoichiometric calibration method for dual energy computed tomography.

    PubMed

    Bourque, Alexandra E; Carrier, Jean-François; Bouchard, Hugo

    2014-04-21

    The accuracy of radiotherapy dose calculation relies crucially on patient composition data. The computed tomography (CT) calibration methods based on the stoichiometric calibration of Schneider et al (1996 Phys. Med. Biol. 41 111-24) are the most reliable to determine electron density (ED) with commercial single energy CT scanners. Along with the recent developments in dual energy CT (DECT) commercial scanners, several methods were published to determine ED and the effective atomic number (EAN) for polyenergetic beams without the need for CT calibration curves. This paper intends to show that with a rigorous definition of the EAN, the stoichiometric calibration method can be successfully adapted to DECT with significant accuracy improvements with respect to the literature without the need for spectrum measurements or empirical beam hardening corrections. Using a theoretical framework of ICRP human tissue compositions and the XCOM photon cross sections database, the revised stoichiometric calibration method yields Hounsfield unit (HU) predictions within less than ±1.3 HU of the theoretical HU calculated from XCOM data averaged over the spectra used (e.g., 80 kVp, 100 kVp, 140 kVp and 140/Sn kVp). A fit of mean excitation energy (I-value) data as a function of EAN is provided in order to determine the ion stopping power of human tissues from ED-EAN measurements. Analysis of the calibration phantom measurements with the Siemens SOMATOM Definition Flash dual source CT scanner shows that the present formalism yields mean absolute errors of (0.3 ± 0.4)% and (1.6 ± 2.0)% on ED and EAN, respectively. For ion therapy, the mean absolute errors for calibrated I-values and proton stopping powers (216 MeV) are (4.1 ± 2.7)% and (0.5 ± 0.4)%, respectively. In all clinical situations studied, the uncertainties in ion ranges in water for therapeutic energies are found to be less than 1.3 mm, 0.7 mm and 0.5 mm for protons, helium and carbon ions respectively, using a

  9. A stoichiometric calibration method for dual energy computed tomography

    NASA Astrophysics Data System (ADS)

    Bourque, Alexandra E.; Carrier, Jean-François; Bouchard, Hugo

    2014-04-01

    The accuracy of radiotherapy dose calculation relies crucially on patient composition data. The computed tomography (CT) calibration methods based on the stoichiometric calibration of Schneider et al (1996 Phys. Med. Biol. 41 111-24) are the most reliable to determine electron density (ED) with commercial single energy CT scanners. Along with the recent developments in dual energy CT (DECT) commercial scanners, several methods were published to determine ED and the effective atomic number (EAN) for polyenergetic beams without the need for CT calibration curves. This paper intends to show that with a rigorous definition of the EAN, the stoichiometric calibration method can be successfully adapted to DECT with significant accuracy improvements with respect to the literature without the need for spectrum measurements or empirical beam hardening corrections. Using a theoretical framework of ICRP human tissue compositions and the XCOM photon cross sections database, the revised stoichiometric calibration method yields Hounsfield unit (HU) predictions within less than ±1.3 HU of the theoretical HU calculated from XCOM data averaged over the spectra used (e.g., 80 kVp, 100 kVp, 140 kVp and 140/Sn kVp). A fit of mean excitation energy (I-value) data as a function of EAN is provided in order to determine the ion stopping power of human tissues from ED-EAN measurements. Analysis of the calibration phantom measurements with the Siemens SOMATOM Definition Flash dual source CT scanner shows that the present formalism yields mean absolute errors of (0.3 ± 0.4)% and (1.6 ± 2.0)% on ED and EAN, respectively. For ion therapy, the mean absolute errors for calibrated I-values and proton stopping powers (216 MeV) are (4.1 ± 2.7)% and (0.5 ± 0.4)%, respectively. In all clinical situations studied, the uncertainties in ion ranges in water for therapeutic energies are found to be less than 1.3 mm, 0.7 mm and 0.5 mm for protons, helium and carbon ions respectively, using a generic

  10. Supplier Selection Using Weighted Utility Additive Method

    NASA Astrophysics Data System (ADS)

    Karande, Prasad; Chakraborty, Shankar

    2015-10-01

    Supplier selection is a multi-criteria decision-making (MCDM) problem which mainly involves evaluating a number of available suppliers according to a set of common criteria for choosing the best one to meet the organizational needs. For any manufacturing or service organization, selecting the right upstream suppliers is a key success factor that will significantly reduce purchasing cost, increase downstream customer satisfaction and improve competitive ability. The past researchers have attempted to solve the supplier selection problem employing different MCDM techniques which involve active participation of the decision makers in the decision-making process. This paper deals with the application of weighted utility additive (WUTA) method for solving supplier selection problems. The WUTA method, an extension of utility additive approach, is based on ordinal regression and consists of building a piece-wise linear additive decision model from a preference structure using linear programming (LP). It adopts preference disaggregation principle and addresses the decision-making activities through operational models which need implicit preferences in the form of a preorder of reference alternatives or a subset of these alternatives present in the process. The preferential preorder provided by the decision maker is used as a restriction of a LP problem, which has its own objective function, minimization of the sum of the errors associated with the ranking of each alternative. Based on a given reference ranking of alternatives, one or more additive utility functions are derived. Using these utility functions, the weighted utilities for individual criterion values are combined into an overall weighted utility for a given alternative. It is observed that WUTA method, having a sound mathematical background, can provide accurate ranking to the candidate suppliers and choose the best one to fulfill the organizational requirements. Two real time examples are illustrated to prove

  11. Calibration Methods Used in Cancer Simulation Models and Suggested Reporting Guidelines

    PubMed Central

    Stout, Natasha K.; Knudsen, Amy B.; Kong, Chung Yin (Joey); McMahon, Pamela M.; Gazelle, G. Scott

    2009-01-01

    Background Increasingly, computer simulation models are used for economic and policy evaluation in cancer prevention and control. A model’s predictions of key outcomes such as screening effectiveness depends on the values of unobservable natural history parameters. Calibration is the process of determining the values of unobservable parameters by constraining model output to replicate observed data. Because there are many approaches for model calibration and little consensus on best practices, we surveyed the literature to catalogue the use and reporting of these methods in cancer simulation models. Methods We conducted a MEDLINE search (1980 through 2006) for articles on cancer screening models and supplemented search results with articles from our personal reference databases. For each article, two authors independently abstracted pre-determined items using a standard form. Data items included cancer site, model type, methods used for determination of unobservable parameter values, and description of any calibration protocol. All authors reached consensus on items of disagreement. Reviews and non-cancer models were excluded. Articles describing analytical models which estimate parameters with statistical approaches (e.g., maximum likelihood) were catalogued separately. Models that included unobservable parameters were analyzed and classified by whether calibration methods were reported and if so, the methods used. Results The review process yielded 154 articles that met our inclusion criteria and of these, we concluded that 131 may have used calibration methods to determine model parameters. Although the term “calibration” was not always used, descriptions of calibration or “model fitting” were found in 50% (n=66) of the articles with an additional 16% (n=21) providing a reference to methods. Calibration target data were identified in nearly all of these articles. Other methodologic details such as the goodness-of-fit metric were discussed in 54% (n=47

  12. Calibration Methods for Air Coupled Antennas - COST Action TU1208

    NASA Astrophysics Data System (ADS)

    Marecos, Vânia; Solla, Mercedes; Fontul, Simona; Pajewski, Lara

    2016-04-01

    This work focuses on the comparison of different methods for calibrating air coupled antennas: Coring, Surface Reflection Method (SRM) and Common Mid-Point (CMP) through the analysis of GPR data collected in a test site with different pavement solutions. Research activities have been carried out during a Short Term Scientific Mission (STSM) funded by the COST (European Cooperation in Science and Technology) Action TU1208 "Civil Engineering Applications of Ground Penetrating Radar" in December 2015. The use of GPR in transport infrastructures represents one of the most significant advances for obtaining continuous data along the road, with the advantage of operation at traffic speed and being a non-destructive technique. Its main application has been the evaluation of layer thickness. For the determination of layer thickness, it is necessary to know the velocity of the signal, which depends on the dielectric constant of the material, and the two-way travel time of the reflected signal that is recorded by the GPR system. The calculation of the dielectric value of the materials can be done using different approaches such as: using fixed values based on experience, laboratory determination of dielectric values, applying the SRM, performing back calculation from ground truth references such as cores and test pits, or using the CMP method. The problem with using ground truth is that it is time consuming, labour intensive and intrusive to traffic, in addition, a drill core is not necessarily representative of the whole surveyed area. Regarding the surface reflection technique, one of the problems is that it only measures the dielectric value from the layer surface and not from the whole layer. Recent works already started to address some of these challenges proposing new approaches for GPR layer thickness measurements using multiple antennas to calculate the average dielectric value of the asphalt layer, taking advantage of significant hardware improvements in GPR

  13. A Calibration Method for Wide-Field Multicolor Photometric Systems

    NASA Astrophysics Data System (ADS)

    Zhou, Xu; Chen, Jiansheng; Xu, Wen; Zhang, Mei; Jiang, Zhaoji; Zheng, Zhongyuan; Zhu, Jin

    1999-07-01

    The purpose of this paper is to present a method to self-calibrate the spectral energy distribution (SED) of objects in a survey based on the fitting of a SED library to observed multicolor photometry. We adopt, for illustrative purposes, the Vilnius and Gunn & Stryker SED libraries. The self-calibration technique can improve the quality of observations which are not taken under perfectly photometric conditions. The more passbands used for the photometry, the better the results. This technique has been applied to the BATC 15 passband CCD survey.

  14. Calibration method for angular measurement of moiré patterns

    NASA Astrophysics Data System (ADS)

    He, Fan; Bai, Jian; Wang, Kaiwei; Hou, Xiyun; Yao, Jiang

    2014-09-01

    In this paper, a high-accuracy calibration method for angular measurement of deformed and curved Moiré patterns, based on template matching algorithm, is presented. We report a feasible and accurate method, based on Talbot interferometry and Moiré deflectometry, to measure long focal-length lenses. Theoretical analysis indicates that the precision of this method is mainly influenced by the angle of Moiré patterns. However, it's difficult to obtain high-accuracy angle of Moiré patterns, since the Moiré patterns derived from experiment are constantly deformed or curved. We demonstrate a method, based on template matching algorithm, to calibrate deformed and curved Moiré patterns, thus their angle can be calculated fast and accurately in sub-pixel domain. Numerical analysis and simulation prove that the method mentioned above demonstrates high precision and stability, and experiment results show that the accuracy of the long focal lengths measurement is improved obviously.

  15. Histogram-Based Calibration Method for Pipeline ADCs.

    PubMed

    Son, Hyeonuk; Jang, Jaewon; Kim, Heetae; Kang, Sungho

    2015-01-01

    Measurement and calibration of an analog-to-digital converter (ADC) using a histogram-based method requires a large volume of data and a long test duration, especially for a high resolution ADC. A fast and accurate calibration method for pipelined ADCs is proposed in this research. The proposed calibration method composes histograms through the outputs of each stage and calculates error sources. The digitized outputs of a stage are influenced directly by the operation of the prior stage, so the results of the histogram provide the information of errors in the prior stage. The composed histograms reduce the required samples and thus calibration time being implemented by simple modules. For 14-bit resolution pipelined ADC, the measured maximum integral non-linearity (INL) is improved from 6.78 to 0.52 LSB, and the spurious-free dynamic range (SFDR) and signal-to-noise-and-distortion ratio (SNDR) are improved from 67.0 to 106.2dB and from 65.6 to 84.8dB, respectively. PMID:26070196

  16. An MLC calibration method using a detector array

    SciTech Connect

    Simon, Thomas A.; Kahler, Darren; Simon, William E.; Fox, Christopher; Li, Jonathan; Palta, Jatinder; Liu, Chihray

    2009-10-15

    Purpose: The authors have developed a quantitative calibration method for a multileaf collimator (MLC) which measures individual leaf positions relative to the MLC backup jaw on an Elekta Synergy linear accelerator. Methods: The method utilizes a commercially available two-axis detector array (Profiler 2; Sun Nuclear Corporation, Melbourne, FL). To calibrate the MLC bank, its backup jaw is positioned at the central axis and the opposing jaw is retracted to create a half-beam configuration. The position of the backup jaws field edge is then measured with the array to obtain what is termed the radiation defined reference line. The positions of the individual leaf ends relative to this reference line are then inferred by the detector response in the leaf end penumbra. Iteratively adjusting and remeasuring the leaf end positions to within specifications completes the calibration. Using the backup jaw as a reference for the leaf end positions is based on three assumptions: (1) The leading edge of an MLC leaf bank is parallel to its backup jaw's leading edge, (2) the backup jaw position is reproducible, and (3) the measured radiation field edge created by each leaf end is representative of that leaf's position. Data from an electronic portal imaging device (EPID) were used in a similar analysis to check the results obtained with the array. Results: The relative leaf end positions measured with the array differed from those measured with the EPID by an average of 0.11 {+-}0.09 mm per leaf. The maximum leaf positional change measured with the Profiler 2 over a 3 month period was 0.51 mm. A leaf positional accuracy of {+-}0.4 mm is easily attainable through the iterative calibration process. The method requires an average of 40 min to measure both leaf banks. Conclusions: This work demonstrates that the Profiler 2 is an effective tool for efficient and quantitative MLC quality assurance and calibration.

  17. A comparison of acromion marker cluster calibration methods for estimating scapular kinematics during upper extremity ergometry.

    PubMed

    Richardson, R Tyler; Nicholson, Kristen F; Rapp, Elizabeth A; Johnston, Therese E; Richards, James G

    2016-05-01

    Accurate measurement of joint kinematics is required to understand the musculoskeletal effects of a therapeutic intervention such as upper extremity (UE) ergometry. Traditional surface-based motion capture is effective for quantifying humerothoracic motion, but scapular kinematics are challenging to obtain. Methods for estimating scapular kinematics include the widely-reported acromion marker cluster (AMC) which utilizes a static calibration between the scapula and the AMC to estimate the orientation of the scapula during motion. Previous literature demonstrates that including additional calibration positions throughout the motion improves AMC accuracy for single plane motions; however this approach has not been assessed for the non-planar shoulder complex motion occurring during UE ergometry. The purpose of this study was to evaluate the accuracy of single, dual, and multiple AMC calibration methods during UE ergometry. The orientations of the UE segments of 13 healthy subjects were recorded with motion capture. Scapular landmarks were palpated at eight evenly-spaced static positions around the 360° cycle. The single AMC method utilized one static calibration position to estimate scapular kinematics for the entire cycle, while the dual and multiple AMC methods used two and four static calibration positions, respectively. Scapulothoracic angles estimated by the three AMC methods were compared with scapulothoracic angles determined by palpation. The multiple AMC method produced the smallest RMS errors and was not significantly different from palpation about any axis. We recommend the multiple AMC method as a practical and accurate way to estimate scapular kinematics during UE ergometry. PMID:26976228

  18. Gravimetric method for in vitro calibration of skin hydration measurements.

    PubMed

    Martinsen, Ørjan G; Grimnes, Sverre; Nilsen, Jon K; Tronstad, Christian; Jang, Wooyoung; Kim, Hongsig; Shin, Kunsoo; Naderi, Majid; Thielmann, Frank

    2008-02-01

    A novel method for in vitro calibration of skin hydration measurements is presented. The method combines gravimetric and electrical measurements and reveals an exponential dependency of measured electrical susceptance to absolute water content in the epidermal stratum corneum. The results also show that absorption of water into the stratum corneum exhibits three different phases with significant differences in absorption time constant. These phases probably correspond to bound, loosely bound, and bulk water. PMID:18270010

  19. Efficient Calibration of Categorical Parameter Distributions using Subspace Methods

    NASA Astrophysics Data System (ADS)

    Khambhammettu, P.; Renard, P.; Doherty, J.

    2014-12-01

    Categorical parameter distributions are common-place in hydrogeological systems consisting of rock-types / aquifer materials with distinct properties, eg: sand channels in a clay matrix. Model calibration is difficult in such systems because the inverse problem is hindered by the discontinuities in the parameter space. In this paper, we present two approaches based on sub-space methods to generate categorical parameter distributions of aquifer parameters that meet calibration constraints (eg:- measured water level data, gradients) while honoring prior geological constraints. In the first approach, the prior geological information and acceptable parameter distributions are encapsulated in a simple object-based model. In the second approach, a Multiple-Point Statistics simulator is used to represent the prior geological information. Sub-space methods in conjunction with dynamic pilot points are then employed to explore the parameter space and determine the parameter combinations that optimally honor geologic and calibration constraints. Using a simple aquifer system, we demonstrate that the new approach is capable of quickly generating multiple multiple parameter distributions that honor both geological and calibration constraints. We also explore the underlying parameter and predictive uncertainty using Null Space Monte Carlo techniques.

  20. Calibration

    NASA Astrophysics Data System (ADS)

    Kunze, Hans-Joachim

    Commercial spectrographic systems are usually supplied with some wave-length calibration, but it is essential that the experimenter performs his own calibration for reliable measurements. A number of sources emitting well-known emission lines are available, and the best values of their wavelengths may be taken from data banks accessible on the internet. Data have been critically evaluated for many decades by the National Institute of Standards and Technology (NIST) of the USA [13], see also p. 3. Special data bases have been established by the astronomy and fusion communities (Appendix B).

  1. A calibration method of the multi-channel imaging lidar

    NASA Astrophysics Data System (ADS)

    Xu, Weiming; Liu, Jun; Shu, Rong

    2014-06-01

    We design a kind of imaging LiDAR with sixteen channels, which consists of a fiber laser source, dual scanning galvanometers, range measurement circuits and information processing circuits etc. The image LiDAR provides sixteen range measurements for one laser shot and the distance accuracy of each channel is about 4cm. This paper provides a calibrate method to correct point cloud images captured with the multi-channel LiDAR. The method needs to construct different slanted planes to cover the imaging field, and establish precise plane equations in the known ground coordinates, then fit planes with point clouds data and calculate correction parameters of all channels through the error model. The image accuracy is better than 5cm processed by this calibration method.

  2. Automatic camera calibration method based on dashed lines

    NASA Astrophysics Data System (ADS)

    Li, Xiuhua; Wang, Guoyou; Liu, Jianguo

    2013-10-01

    We present a new method for full-automatic calibration of traffic cameras using the end points on dashed lines. Our approach uses the improved RANSAC method with the help of pixels transverse projection to detect the dashed lines and end points on them. Then combining analysis of the geometric relationship between the camera and road coordinate systems, we construct a road model to fit the end points. Finally using two-dimension calibration method we can convert pixels in image to meters along the ground truth lane. On a large number of experiments exhibiting a variety of conditions, our approach performs well, achieving less than 5% error in measuring test lengths in all cases.

  3. a Modified Method for Polarimetric SAR Calibration Algorithm

    NASA Astrophysics Data System (ADS)

    Liao, L.; Li, P.; Yang, J.

    2013-07-01

    Present fully polarimetric synthetic aperture radar (SAR) systems often update calibration techniques to further enhance the accuracy to the polarimetric data. In this paper, we propose a modified method to estimate the value of crosstalk based on the corrected observed value. Since Ainsworth calibration algorithm firstly set the value of k to be one. And the value of k relates to the copolarization channel imbalance .We consider the effects of value of k and analyze it. Through comparison to crosstalk results between the stimulated parameters and the estimated parameters, we assume high co-polarization channel imbalance will be obviously to affect crosstalk results. Then, used covariance observation value of the initial value of k rewrites the model to solve related parameters. And crosstalk parameter is calculated by the same iterative method. To verify the effect of the modified calibration method, this letter compares the accuracy of the two methods using the simulated polarimetric SAR data and Chinese airborne X-band polarimetric SAR data. The results confirm that the modified method tends to get more accurate crosstalk results.

  4. AVIRIS calibration using the cloud-shadow method

    NASA Technical Reports Server (NTRS)

    Carder, K. L.; Reinersman, P.; Chen, R. F.

    1993-01-01

    More than 90 percent of the signal at an ocean-viewing, satellite sensor is due to the atmosphere, so a 5 percent sensor-calibration error viewing a target that contributes but 10 percent of the signal received at the sensor may result in a target-reflectance error of more than 50 percent. Since prelaunch calibration accuracies of 5 percent are typical of space-sensor requirements, recalibration of the sensor using ground-base methods is required for low-signal target. Known target reflectance or water-leaving radiance spectra and atmospheric correction parameters are required. In this article we describe an atmospheric-correction method that uses cloud shadowed pixels in combination with pixels in a neighborhood region of similar optical properties to remove atmospheric effects from ocean scenes. These neighboring pixels can then be used as known reflectance targets for validation of the sensor calibration and atmospheric correction. The method uses the difference between water-leaving radiance values for these two regions. This allows nearly identical optical contributions to the two signals (e.g., path radiance and Fresnel-reflected skylight) to be removed, leaving mostly solar photons backscattered from beneath the sea to dominate the residual signal. Normalization by incident solar irradiance reaching the sea surface provides the remote-sensing reflectance of the ocean at the location of the neighbor region.

  5. An alternative calibration method for counting P-32 reactor monitors

    SciTech Connect

    Quirk, T.J.; Vehar, D.W.

    2011-07-01

    Radioactivation of sulfur is a common technique used to measure fast neutron fluences in test and research reactors. Elemental sulfur can be pressed into pellets and used as monitors. The {sup 32}S(n, p) {sup 32}P reaction has a practical threshold of about 3 MeV and its cross section and associated uncertainties are well characterized [1]. The product {sup 32P} emits a beta particle with a maximum energy of 1710 keV [2]. This energetic beta particle allows pellets to be counted intact. ASTM Standard Test Method for Measuring Reaction Rates and Fast-Neutron Fluences by Radioactivation of Sulfur-32 (E265) [3] details a method of calibration for counting systems and subsequent analysis of results. This method requires irradiation of sulfur monitors in a fast-neutron field whose spectrum and intensity are well known. The resultant decay-corrected count rate is then correlated to the known fast neutron fluence. The Radiation Metrology Laboratory (RML) at Sandia has traditionally performed calibration irradiations of sulfur pellets using the {sup 252}Cf spontaneous fission neutron source at the National Inst. of Standards and Technology (NIST) [4] as a transfer standard. However, decay has reduced the intensity of NIST's source; thus lowering the practical upper limits of available fluence. As of May 2010, neutron emission rates have decayed to approximately 3 e8 n/s. In practice, this degradation of capabilities precludes calibrations at the highest fluence levels produced at test reactors and limits the useful range of count rates that can be measured. Furthermore, the reduced availability of replacement {sup 252}Cf threatens the long-term viability of the NIST {sup 252}Cf facility for sulfur pellet calibrations. In lieu of correlating count rate to neutron fluence in a reference field the total quantity of {sup 32}P produced in a pellet can be determined by absolute counting methods. This offers an attractive alternative to extended {sup 252}Cf exposures because it

  6. Accurate camera calibration method specialized for virtual studios

    NASA Astrophysics Data System (ADS)

    Okubo, Hidehiko; Yamanouchi, Yuko; Mitsumine, Hideki; Fukaya, Takashi; Inoue, Seiki

    2008-02-01

    Virtual studio is a popular technology for TV programs, that makes possible to synchronize computer graphics (CG) to realshot image in camera motion. Normally, the geometrical matching accuracy between CG and realshot image is not expected so much on real-time system, we sometimes compromise on directions, not to come out the problem. So we developed the hybrid camera calibration method and CG generating system to achieve the accurate geometrical matching of CG and realshot on virtual studio. Our calibration method is intended for the camera system on platform and tripod with rotary encoder, that can measure pan/tilt angles. To solve the camera model and initial pose, we enhanced the bundle adjustment algorithm to fit the camera model, using pan/tilt data as known parameters, and optimizing all other parameters invariant against pan/tilt value. This initialization yields high accurate camera position and orientation consistent with any pan/tilt values. Also we created CG generator implemented the lens distortion function with GPU programming. By applying the lens distortion parameters obtained by camera calibration process, we could get fair compositing results.

  7. New Method for Calibration for Hyperspectral Pushbroom Imaging Systems

    NASA Technical Reports Server (NTRS)

    Ryan, Robert; Olive, Dan; ONeal, Duane; Schere, Chris; Nixon, Thomas; May, Chengye; Ryan, Jim; Stanley, Tom; Witcher, Kern

    1999-01-01

    A new, easy-to-implement approach for achieving highly accurate spectral and radiometric calibration of array-based, hyperspectral pushbroom imagers is presented in this paper. The equivalence of the plane of the exit port of an integrating sphere to a Lambertian surface is utilized to provide a field-filling radiance source for the imager. Several different continuous wave lasers of various wavelengths and a quartz-tungsten-halogen lamp internally illuminate the sphere. The imager is positioned to "stare" into the port, and the resultant data cube is analyzed to determine wavelength calibrations, spectral widths of channels, radiometric characteristics, and signal-to-noise ratio, as well as an estimate of signal-to-noise performance in the field. The "smile" (geometric distortion of spectra) of the system can be quickly ascertained using this method. As the price and availability of solid state laser sources improve, this technique could gain wide acceptance.

  8. Sum of ranking differences (SRD) to ensemble multivariate calibration model merits for tuning parameter selection and comparing calibration methods.

    PubMed

    Kalivas, John H; Héberger, Károly; Andries, Erik

    2015-04-15

    Most multivariate calibration methods require selection of tuning parameters, such as partial least squares (PLS) or the Tikhonov regularization variant ridge regression (RR). Tuning parameter values determine the direction and magnitude of respective model vectors thereby setting the resultant predication abilities of the model vectors. Simultaneously, tuning parameter values establish the corresponding bias/variance and the underlying selectivity/sensitivity tradeoffs. Selection of the final tuning parameter is often accomplished through some form of cross-validation and the resultant root mean square error of cross-validation (RMSECV) values are evaluated. However, selection of a "good" tuning parameter with this one model evaluation merit is almost impossible. Including additional model merits assists tuning parameter selection to provide better balanced models as well as allowing for a reasonable comparison between calibration methods. Using multiple merits requires decisions to be made on how to combine and weight the merits into an information criterion. An abundance of options are possible. Presented in this paper is the sum of ranking differences (SRD) to ensemble a collection of model evaluation merits varying across tuning parameters. It is shown that the SRD consensus ranking of model tuning parameters allows automatic selection of the final model, or a collection of models if so desired. Essentially, the user's preference for the degree of balance between bias and variance ultimately decides the merits used in SRD and hence, the tuning parameter values ranked lowest by SRD for automatic selection. The SRD process is also shown to allow simultaneous comparison of different calibration methods for a particular data set in conjunction with tuning parameter selection. Because SRD evaluates consistency across multiple merits, decisions on how to combine and weight merits are avoided. To demonstrate the utility of SRD, a near infrared spectral data set and a

  9. Self-calibration method of two-dimensional grid plate

    NASA Astrophysics Data System (ADS)

    Ding, Guoqing; Chen, Xin; Wang, Lihua; Lei, Lihua; Li, Yuan

    2011-12-01

    A two-dimensional grid plate can offer an X-Y position standard where grids are aligned orthogonal to each other. It is important to ensure the positional accuracy of the grid plate when the grid plate is used to calibrate planar movement systems, such as vision measuring machines and scanning probe microscopes. Existing algorithms for self-calibration employ the discrete Fourier transform, which is complicated and has poor noise suppression capability. We have developed an algorithm that can achieve exact self-calibration for a two-dimensional grid plate using the least squares method when there is no random noise. In the presence of random noise, the algorithm still presents an excellent capability for noise suppression. As an extension of the classic three-location measurement, the algorithm can be applied to four- or five-location measurements, which reduce measurement uncertainties. The error propagation characteristic of the random errors has been investigated in the case of different measurement strategies. According to the simulation results, the mean error propagation ratios are less than 1 when the array size of the grid plate is less than 32×32. Finally, the influence of the scale errors of the planar movement system is discussed.

  10. Method of Calibration for a Large Cathetometer System

    NASA Technical Reports Server (NTRS)

    Toland, Ronald

    2004-01-01

    A method of calibration has been devised for a pair of mutually orthogonal two-axis cathetometers that, when used together, yield measurements of three-dimensional positions of objects mounted on an optical bench. Each cathetometer has a horizontal travel of 1.8 m and a vertical travel of 1.2 m. The cathetometers are required to measure X, Y, and Z coordinates (see figure) to within plus or minus 0.005 in. (plus or minus 0.127 mm). Each cathetometer consists of an alignment telescope on a platform mounted on a two-dimensional translation stage. The knowledge required for calibration of each cathetometer is (1) the two-dimensional position of the cathetometer platform as a function of the electronic readouts of position encoders on the translation stage and (2) the amount of any angular misalignment (roll, pitch, and/or yaw) of the cathetometer platform as a function of the two-dimensional coordinates or the position-encoder readouts. By use of three equations derived from the applicable trigonometric relationships, the calibrated X, Y, and Z coordinates can be computed from the raw encoder readouts.

  11. A geometric calibration method for cone beam CT systems

    SciTech Connect

    Yang, Kai; Kwan, Alexander L. C.; Miller, DeWitt F.; Boone, John M.

    2006-06-15

    Cone beam CT systems are being deployed in large numbers for small animal imaging, dental imaging, and other specialty applications. A new high-precision method for cone beam CT system calibration is presented in this paper. It uses multiple projection images acquired from rotating point-like objects (metal ball bearings) and the angle information generated from the rotating gantry system is also used. It is assumed that the whole system has a mechanically stable rotation center and that the detector does not have severe out-of-plane rotation (<2 deg.). Simple geometrical relationships between the orbital paths of individual BBs and five system parameters were derived. Computer simulations were employed to validate the accuracy of this method in the presence of noise. Equal or higher accuracy was achieved compared with previous methods. This method was implemented for the geometrical calibration of both a micro CT scanner and a breast CT scanner. The reconstructed tomographic images demonstrated that the proposed method is robust and easy to implement with high precision.

  12. Calibrated breast density methods for full field digital mammography: A system for serial quality control and inter-system generalization

    PubMed Central

    Lu, B.; Smallwood, A. M.; Sellers, T. A.; Drukteinis, J. S.; Heine, J. J.

    2015-01-01

    Purpose: The authors are developing a system for calibrated breast density measurements using full field digital mammography (FFDM). Breast tissue equivalent (BTE) phantom images are used to establish baseline (BL) calibration curves at time zero. For a given FFDM unit, the full BL dataset is comprised of approximately 160 phantom images, acquired prior to calibrating prospective patient mammograms. BL curves are monitored serially to ensure they produce accurate calibration and require updating when calibration accuracy degrades beyond an acceptable tolerance, rather than acquiring full BL datasets repeatedly. BL updating is a special case of generalizing calibration datasets across FFDM units, referred to as cross-calibration. Serial monitoring, BL updating, and cross-calibration techniques were developed and evaluated. Methods: BL curves were established for three Hologic Selenia FFDM units at time zero. In addition, one set of serial phantom images, comprised of equal proportions of adipose and fibroglandular BTE materials (50/50 compositions) of a fixed height, was acquired biweekly and monitored with the cumulative sum (Cusum) technique. These 50/50 composition images were used to update the BL curves when the calibration accuracy degraded beyond a preset tolerance of ±4 standardized units. A second set of serial images, comprised of a wide-range of BTE compositions, was acquired biweekly to evaluate serial monitoring, BL updating, and cross-calibration techniques. Results: Calibration accuracy can degrade serially and is a function of acquisition technique and phantom height. The authors demonstrated that all heights could be monitored simultaneously while acquiring images of a 50/50 phantom with a fixed height for each acquisition technique biweekly, translating into approximately 16 image acquisitions biweekly per FFDM unit. The same serial images are sufficient for serial monitoring, BL updating, and cross-calibration. Serial calibration accuracy was

  13. The cryogenic balance design and balance calibration methods

    NASA Astrophysics Data System (ADS)

    Ewald, B.; Polanski, L.; Graewe, E.

    1992-07-01

    The current status of a program aimed at the development of a cryogenic balance for the European Transonic Wind Tunnel is reviewed. In particular, attention is given to the cryogenic balance design philosophy, mechanical balance design, reliability and accuracy, cryogenic balance calibration concept, and the concept of an automatic calibration machine. It is shown that the use of the automatic calibration machine will improve the accuracy of calibration while reducing the man power and time required for balance calibration.

  14. Various methods and developments for calibrating seismological sensors at EOST

    NASA Astrophysics Data System (ADS)

    JUND, H.; Bès de Berc, M.; Thore, J.

    2013-12-01

    Calibrating seismic sensors is crucial for knowing the quality of the sensor and generating precise dataless files. We present here three calibration methods that we have developed for the short period and broad band sensors included in the temporary and permanent seismic networks in France. First, in the case of a short-period sensor with no electronics and calibration coil, we inject a sine wave signal into the signal coil. After locking the sensor mass, we first connect a voltage generator of signal waves and a series resistor to the coil. Then, a sinusoidal signal is sent to the sensor signal coil output. Both the voltage at the terminal of the resistor, which gives an image of the intensity entering the signal coil, and the voltage at the terminal of the signal coil are measured. The frequency of the generator then varies in order to find a phase shift between both signals of π/2. The output frequency of the generator corresponds to the image of the natural frequency of the sensor. Second, in the case of all types of sensors provided with a calibration coil, we inject different signals into the calibration coil. We usually apply two signals: a step signal and a sweep (or wobble) signal. A step signal into the calibration coil is equivalent to a Dirac excitation in derived acceleration. The response to this Dirac gives the transfer function of the signal coil, derived two times and without absolute gain. We developed a field-module allowing us to always apply the same excitation to various models of seismometers, in order to compare the results from several instruments previously installed on field. A wobble signal is a signal whose frequency varies. By varying the frequency of the input signal around the sensor's natural frequency, we obtain an immediate response of the sensor in acceleration. This method is particularly suitable in order to avoid any disturbances which may modify the signal of a permanent station. Finally, for the determination of absolute

  15. Simple method for absolute calibration of geophones, seismometers, and other inertial vibration sensors

    SciTech Connect

    Kann, Frank van; Winterflood, John

    2005-03-01

    A simple but powerful method is presented for calibrating geophones, seismometers, and other inertial vibration sensors, including passive accelerometers. The method requires no cumbersome or expensive fixtures such as shaker platforms and can be performed using a standard instrument commonly available in the field. An absolute calibration is obtained using the reciprocity property of the device, based on the standard mathematical model for such inertial sensors. It requires only simple electrical measurement of the impedance of the sensor as a function of frequency to determine the parameters of the model and hence the sensitivity function. The method is particularly convenient if one of these parameters, namely the suspended mass is known. In this case, no additional mechanical apparatus is required and only a single set of impedance measurements yields the desired calibration function. Moreover, this measurement can be made with the device in situ. However, the novel and most powerful aspect of the method is its ability to accurately determine the effective suspended mass. For this, the impedance measurement is made with the device hanging from a simple spring or flexible cord (depending on the orientation of its sensitive axis). To complete the calibration, the device is weighed to determine its total mass. All the required calibration parameters, including the suspended mass, are then determined from a least-squares fit to the impedance as a function of frequency. A demonstration using both a 4.5 Hz geophone and a 1 Hz seismometer shows that the method can yield accurate absolute calibrations with an error of 0.1% or better, assuming no a priori knowledge of any parameters.

  16. A quick telemanipulator calibration and repeatability method with applications

    SciTech Connect

    Jansen, J.F.; Haley, D.C.

    1994-09-01

    This paper will present a methodology that was used to calibrate and measure the repeatability of two telemanipulators at Oak Ridge National Laboratory. The global accuracy of the method was 0.05 in. ({approx_equal} 1.3 mm), and the orientation accuracy was approximately 6 min ({approx_equal} 0.002 rads). For most teleoperator systems, these accuracies are more than adequate because of the construction of the mechanism and sensor capabilities (e.g., typically 12 bits of resolution). Although industrial robots require accuracies of about 0.05 mm or better, telemanipulators do not.

  17. Method of calibrating a fluid-level measurement system

    NASA Technical Reports Server (NTRS)

    Woodard, Stanley E. (Inventor); Taylor, Bryant D. (Inventor)

    2010-01-01

    A method of calibrating a fluid-level measurement system is provided. A first response of the system is recorded when the system's sensor(s) is (are) not in contact with a fluid of interest. A second response of the system is recorded when the system's sensor(s) is (are) fully immersed in the fluid of interest. Using the first and second responses, a plurality of expected responses of the system's sensor(s) is (are) generated for a corresponding plurality of levels of immersion of the sensor(s) in the fluid of interest.

  18. Method of making self-calibrated displacement measurements

    DOEpatents

    Pedersen, Herbert N.

    1977-01-01

    A method for monitoring the displacement of an object having an acoustically reflective surface at least partially submerged in an acoustically conductive medium. The reflective surface is designed to have a stepped interface responsive to an incident acoustic pulse to provide separate discrete reflected pulses to a receiving transducer. The difference in the time of flight of the reflected acoustic signals corresponds to the known step height and the time of travel of the signals to the receiving transducer provides a measure of the displacement of the object. Accordingly, the reference step length enables simultaneous calibration of each displacement measurement.

  19. Spinning disk calibration method and apparatus for laser Doppler velocimeter

    NASA Technical Reports Server (NTRS)

    Snyder, P. K. (Inventor)

    1986-01-01

    A method and apparatus for calibrating laser Doppler velocimeters having one or more intersecting beam pairs are described. These velocimeters measure fluid velocity by observing the light scattered by particles in the fluid stream. Moving fluid particulates are simulated by fine taut wires that are radially mounted on a disk that is rotated at a known velocity. The laser beam intersection locus is first aimed at the very center of the disk and then the disk is translated so that the locus is swept by the rotating wires. The radial distance traversed is precisely measured so that the velocity of the wires (pseudo particles) may be calculated.

  20. Investigation of factors affecting the heater wire method of calibrating fine wire thermocouples

    NASA Technical Reports Server (NTRS)

    Keshock, E. G.

    1972-01-01

    An analytical investigation was made of a transient method of calibrating fine wire thermocouples. The system consisted of a 10 mil diameter standard thermocouple (Pt, Pt-13% Rh) and an 0.8 mil diameter chromel-alumel thermocouple attached to a 20 mil diameter electrically heated platinum wire. The calibration procedure consisted of electrically heating the wire to approximately 2500 F within about a seven-second period in an environment approximating atmospheric conditions at 120,000 feet. Rapid periodic readout of the standard and fine wire thermocouple signals permitted a comparison of the two temperature indications. An analysis was performed which indicated that the temperature distortion at the heater wire produced by the thermocouple junctions appears to be of negligible magnitude. Consequently, the calibration technique appears to be basically sound, although several practical changes which appear desirable are presented and discussed. Additional investigation is warranted to evaluate radiation effects and transient response characteristics.

  1. Waveform correlation methods for identifying populations of calibration events

    SciTech Connect

    Harris, D.B.

    1997-07-01

    An approach for systematically screening large volumes of continuous data for repetitive events identified as mining explosions on basis of temporal and amplitude population characteristics. The method extends event clustering through waveform correlation with a new source-region-specific detector. The new signal subspace detector generalizes the matched filter and can be used to increase the number of events associated with a given cluster, thereby increasing the reliability of diagnostic cluster population characteristics. The method can be applied to obtain bootstrap ground truth explosion waveforms for testing discriminants, where actual ground truth is absent. The same events, if associated with to a particular mine, may help calibrate velocity models. The method may also assist earthquake hazard risk assessment by providing what amounts to blasting logs for identified mines. The cluster event lists can be reconciled against earthquake catalogs to screen explosions, otherwise hard to identify from the catalogs.

  2. A least squares method for CVT calibration in a RLC capacitor discharge circuit.

    SciTech Connect

    Yao, Stephen E.; Dickey, Fred McCartney; Pecak, Sara North

    2003-11-01

    In many applications, the ability to monitor the output of a capacitive discharge circuit is imperative to ensuring the reliability and accuracy of the unit. This monitoring is commonly accomplished with the use of a Current Viewing Transformer (CVT). In order to calibrate the CVT, the circuit is assembled with a Current Viewing Transformer (CVR) in addition to the CVT and the peak outputs compared. However, difficulties encountered with the use of CVRs make it desirable to eliminate the use of the CVR from the calibration process. This report describes a method for determining the calibration factor between the current throughput and the CVT voltage output in a capacitive discharge unit from the CVT ringdown data and values of initial voltage and capacitance of the circuit. Previous linear RLC fitting work for determining R, L, and C is adapted to return values of R, L, and the calibration factor, k. Separate solutions for underdamped and overdamped cases are presented and implemented on real circuit data using MathCad software with positive results. This technique may also offer a unique approach to self calibration of current measuring devices.

  3. A method of calibrating wind velocity sensors with a modified gas flow calibrator

    NASA Technical Reports Server (NTRS)

    Stump, H. P.

    1978-01-01

    A procedure was described for calibrating air velocity sensors in the exhaust flow of a gas flow calibrator. The average velocity in the test section located at the calibrator exhaust was verified from the mass flow rate accurately measured by the calibrator's precision sonic nozzles. Air at elevated pressures flowed through a series of screens, diameter changes, and flow straighteners, resulting in a smooth flow through the open test section. The modified system generated air velocities of 2 to 90 meters per second with an uncertainty of about two percent for speeds below 15 meters per second and four percent for the higher speeds. Wind tunnel data correlated well with that taken in the flow calibrator.

  4. Calibration Methods for a 3D Triangulation Based Camera

    NASA Astrophysics Data System (ADS)

    Schulz, Ulrike; Böhnke, Kay

    A sensor in a camera takes a gray level image (1536 x 512 pixels), which is reflected by a reference body. The reference body is illuminated by a linear laser line. This gray level image can be used for a 3D calibration. The following paper describes how a calibration program calculates the calibration factors. The calibration factors serve to determine the size of an unknown reference body.

  5. Calibration of the DRASTIC ground water vulnerability mapping method

    USGS Publications Warehouse

    Rupert, M.G.

    2001-01-01

    Ground water vulnerability maps developed using the DRASTIC method have been produced in many parts of the world. Comparisons of those maps with actual ground water quality data have shown that the DRASTIC method is typically a poor predictor of ground water contamination. This study significantly improved the effectiveness of a modified DRASTIC ground water vulnerability map by calibrating the point rating schemes to actual ground water quality data by using nonparametric statistical techniques and a geographic information system. Calibration was performed by comparing data on nitrite plus nitrate as nitrogen (NO2 + NO3-N) concentrations in ground water to land-use, soils, and depth to first-encountered ground water data. These comparisons showed clear statistical differences between NO2 + NO3-N concentrations and the various categories. Ground water probability point ratings for NO2 + NO3-N contamination were developed from the results of these comparisons, and a probability map was produced. This ground water probability map was then correlated with an independent set of NO2 + NO3-N data to demonstrate its effectiveness in predicting elevated NO2 + NO3-N concentrations in ground water. This correlation demonstrated that the probability map was effective, but a vulnerability map produced with the uncalibrated DRASTIC method in the same area and using the same data layers was not effective. Considerable time and expense have been outlaid to develop ground water vulnerability maps with the DRASTIC method. This study demonstrates a cost-effective method to improve and verify the effectiveness of ground water vulnerability maps.

  6. Method for Ground-to-Satellite Laser Calibration System

    NASA Technical Reports Server (NTRS)

    Lukashin, Constantine (Inventor); Wielicki, Bruce A. (Inventor)

    2015-01-01

    The present invention comprises an approach for calibrating the sensitivity to polarization, optics degradation, spectral and stray light response functions of instruments on orbit. The concept is based on using an accurate ground-based laser system, Ground-to-Space Laser Calibration (GSLC), transmitting laser light to instrument on orbit during nighttime substantially clear-sky conditions. To minimize atmospheric contribution to the calibration uncertainty the calibration cycles should be performed in short time intervals, and all required measurements are designed to be relative. The calibration cycles involve ground operations with laser beam polarization and wavelength changes.

  7. Method for Ground-to-Space Laser Calibration System

    NASA Technical Reports Server (NTRS)

    Lukashin, Constantine (Inventor); Wielicki, Bruce A. (Inventor)

    2014-01-01

    The present invention comprises an approach for calibrating the sensitivity to polarization, optics degradation, spectral and stray light response functions of instruments on orbit. The concept is based on using an accurate ground-based laser system, Ground-to-Space Laser Calibration (GSLC), transmitting laser light to instrument on orbit during nighttime substantially clear-sky conditions. To minimize atmospheric contribution to the calibration uncertainty the calibration cycles should be performed in short time intervals, and all required measurements are designed to be relative. The calibration cycles involve ground operations with laser beam polarization and wavelength changes.

  8. A review of some radiometric calibration problems and methods

    NASA Technical Reports Server (NTRS)

    Slater, P. N.

    1984-01-01

    The in-flight radiometric calibration instrumentation and procedures of the Landsat Thematic Mapper and the high-resolution visible-range instruments of SPOT are illustrated with drawings and diagrams, characterized, and compared. Problems encountered in the laboratory calibration process, minimizing the temporal instability of the systems, identifying anomalies in the electronics in flight, and rechecking the calibration are examined, and it is pointed out that the stability of the calibration systems is less than that of the instruments themselves. The use of carefully measured ground-site data and atmospheric parameters in combination with radiative-transfer models for periodic calibration is recommended.

  9. Advances in calibration methods for micro- and nanoscale surfaces

    NASA Astrophysics Data System (ADS)

    Leach, R. K.; Giusca, C. L.; Coupland, J. M.

    2012-04-01

    Optical surface topography measuring instrument manufacturers often quote accuracies of the order of nanometres and claim that the instruments can reliably measure a range of surfaces with structures on the micro- to nanoscale. However, for many years there has been debate about the interpretation of the data from optical surface topography measuring instruments. Optical artefacts in the output data and a lack of a calibration infrastructure mean that it can be difficult to get optical instruments to agree with contact stylus instruments. In this paper, the current situation with areal surface topography measurements is discussed along with the ISO specification standards that are in draft form. An infrastructure is discussed whereby the ISO-defined metrological characteristics of optical instruments can be determined, but these characteristics do not allow the instrument to measure complex surfaces. Current research into methods for determining the transfer function of optical instruments is reviewed, which will allow the calibration of optical instruments to measure complex surfaces, at least in the case of weak scattering. The ability of some optical instruments to measure outside the spatial bandwidth limitation of the numerical aperture is presented and some general outlook for future work given.

  10. Transmission and division of total optical depth method: A universal calibration method for Sun photometric measurements

    NASA Astrophysics Data System (ADS)

    Zhang, Ming; Gong, Wei; Ma, Yingying; Wang, Lunche; Chen, Zhongyong

    2016-03-01

    Sun photometric measurements, which provide accurate and timely information on atmospheric components such as aerosols, clouds, and gases are important to climate research. For regions with heavy and variable aerosol loading, the traditional Langley plot method cannot be applied for Sun photometric instrument calibration, as almost no suitable prolonged periods with stable atmosphere and low-aerosol loading occurs. An improved calibration method, namely, the transmission and division of total optical depth method, is proposed in this study. Atmospheric total optical depth variation information obtained via other methods is transmitted, and period groups with similar atmospheric extinction effects are selected for Langley regression. This method is validated through calibration of a multifilter rotating shadowband radiometer under heavy aerosol-loading conditions. The obtained aerosol optical depth (AOD) compares well with the interpolated AOD from a Cimel Sun-sky radiometer.

  11. Flight Test Results of an Angle of Attack and Angle of Sideslip Calibration Method Using Output-Error Optimization

    NASA Technical Reports Server (NTRS)

    Siu, Marie-Michele; Martos, Borja; Foster, John V.

    2013-01-01

    As part of a joint partnership between the NASA Aviation Safety Program (AvSP) and the University of Tennessee Space Institute (UTSI), research on advanced air data calibration methods has been in progress. This research was initiated to expand a novel pitot-static calibration method that was developed to allow rapid in-flight calibration for the NASA Airborne Subscale Transport Aircraft Research (AirSTAR) facility. This approach uses Global Positioning System (GPS) technology coupled with modern system identification methods that rapidly computes optimal pressure error models over a range of airspeed with defined confidence bounds. Subscale flight tests demonstrated small 2-s error bounds with significant reduction in test time compared to other methods. Recent UTSI full scale flight tests have shown airspeed calibrations with the same accuracy or better as the Federal Aviation Administration (FAA) accepted GPS 'four-leg' method in a smaller test area and in less time. The current research was motivated by the desire to extend this method for inflight calibration of angle of attack (AOA) and angle of sideslip (AOS) flow vanes. An instrumented Piper Saratoga research aircraft from the UTSI was used to collect the flight test data and evaluate flight test maneuvers. Results showed that the output-error approach produces good results for flow vane calibration. In addition, maneuvers for pitot-static and flow vane calibration can be integrated to enable simultaneous and efficient testing of each system.

  12. Methods of calibration in the direct analysis of solid samples by electrothermal atomic absorption spectrometry

    NASA Astrophysics Data System (ADS)

    Berglund, M.; Baxter, D. C.

    1992-12-01

    One of the major problems involved in the direct analysis of solid samples by electrothermal atomic absorption spectrometry (ETAAS) lies in the calibration step because non-spectral interference effects are often pronounced. Three standardization techniques have been described and used in solid sampling-ETAAS: (i) standard additions method; (ii) calibration relative to a certified reference material; and (iii) calibration curve technique. However, an adequate statistical evaluation of the uncertainty in the analyte concentration in the solid sample is most frequently neglected, and reported errors may be seriously underestimated. This can be attributed directly to the complexity of the statistical expressions required to accurately account for errors in each of the calibration techniques mentioned above, and the general lack of relevant reference literature. The object of this work has been to develop a computer package which will perform the necessary statistical analyses of solid sampling-ETAAS data; the result is the program "SOLIDS" described here in the form of an electronic publication in Spectrochimica Acta Electronica, the electronic section of Spectrochimica Acta Part B. The program could also be useful in other analytical fields where similar calibration methods are used. The hard copy text, outlining the calibration models and their associated errors, is accompanied by a diskette containing the program, some data files, and a manual. Use of the program is exemplified in the text, with some of the data files discussed included on the diskette which, together with the manual, should enable the reader to become familiarized with the operation of the program, and the results generated.

  13. Method for in-situ calibration of electrophoretic analysis systems

    DOEpatents

    Liu, Changsheng; Zhao, Hequan

    2005-05-08

    An electrophoretic system having a plurality of separation lanes is provided with an automatic calibration feature in which each lane is separately calibrated. For each lane, the calibration coefficients map a spectrum of received channel intensities onto values reflective of the relative likelihood of each of a plurality of dyes being present. Individual peaks, reflective of the influence of a single dye, are isolated from among the various sets of detected light intensity spectra, and these can be used to both detect the number of dye components present, and also to establish exemplary vectors for the calibration coefficients which may then be clustered and further processed to arrive at a calibration matrix for the system. The system of the present invention thus permits one to use different dye sets to tag DNA nucleotides in samples which migrate in separate lanes, and also allows for in-situ calibration with new, previously unused dye sets.

  14. Simple method for calibrating omnidirectional stereo with multiple cameras

    NASA Astrophysics Data System (ADS)

    Ha, Jong-Eun; Choi, I.-Sak

    2011-04-01

    Cameras can give useful information for the autonomous navigation of a mobile robot. Typically, one or two cameras are used for this task. Recently, an omnidirectional stereo vision system that can cover the whole surrounding environment of a mobile robot is adopted. They usually adopt a mirror that cannot offer uniform spatial resolution. In this paper, we deal with an omnidirectional stereo system which consists of eight cameras where each two vertical cameras constitute one stereo system. Camera calibration is the first necessary step to obtain 3D information. Calibration using a planar pattern requires many images acquired under different poses so it is a tedious step to calibrate all eight cameras. In this paper, we present a simple calibration procedure using a cubic-type calibration structure that surrounds the omnidirectional stereo system. We can calibrate all the cameras on an omnidirectional stereo system in just one shot.

  15. Medical color displays and their color calibration: investigations of various calibration methods, tools, and potential improvement in color difference ΔE

    NASA Astrophysics Data System (ADS)

    Roehrig, Hans; Hashmi, Syed F.; Dallas, William J.; Krupinski, Elizabeth A.; Rehm, Kelly; Fan, Jiahua

    2010-08-01

    that those calibration parameters did conform, with the help of a state of the art Spectroradiometer, PR670. As a result of this addition of the PR670, and also an in-house developed method of profiling and characterization, it appears that there was much improvement in ΔE, the color difference.

  16. One step geometrical calibration method for optical coherence tomography

    NASA Astrophysics Data System (ADS)

    Díaz Díaz, Jesús; Stritzel, Jenny; Rahlves, Maik; Majdani, Omid; Reithmeier, Eduard; Ortmaier, Tobias; Roth, Bernhard

    2016-01-01

    We present a novel one-step calibration methodology for geometrical distortion correction for optical coherence tomography (OCT). A calibration standard especially designed for OCT is introduced, which consists of an array of inverse pyramidal structures. The use of multiple landmarks situated on four different height levels on the pyramids allow performing a 3D geometrical calibration. The calibration procedure itself is based on a parametric model of the OCT beam propagation. It is validated by experimental results and enables the reduction of systematic errors by more than one order of magnitude. In future, our results can improve OCT image reconstruction and interpretation for medical applications such as real time monitoring of surgery.

  17. Combined non-contact coordinate measurement system and calibration method

    NASA Astrophysics Data System (ADS)

    Fan, Yiyan; Zhao, Bin

    2015-07-01

    A combined non-contact measurement system comprising attitude angle sensor, angle encoder, laser rangefinder, and total station is adopted to measure the spatial coordinate of the hidden zones in large-scale space. The laser from the total station is aimed at the optical system of the attitude angle sensor to obtain the spatial coordinate and the spatial attitude angles. Then, the angle encoder driven by a stepping motor is rotated to drive the laser rangefinder to direct at the measured point. This approach is used to obtain the distance from the rangefinder to the measured point and the angle of the angle encoder. Finally, the spatial coordinates of the measured point can be calculated by using these measured parameters. For the measurement system, we propose a weighted least squares (WLS) calibration method, in which weights are determined for the angular distribution density. Experimental results show that the measurement system could expand the scale and achieve reliable precision during combined measurement and the measurement error of the weighted least squares method is less than that of the ordinary least square (OLS) method.

  18. Study for external calibration method for cloud profiling radar on EarthCARE

    NASA Astrophysics Data System (ADS)

    Horie, Hiroaki; Kimura, Toshiyoshi; Okada, Kazuyuki; Ohno, Yuichi; Sato, Kenji; Kumagai, Hiroshi

    2008-10-01

    EarthCARE mission has objectives to reveal aerosol and cloud interaction and to reveal relationships with radiation budget. For this purpose, the EarthCARE satellite has four instruments, which are Atmospheric LIDAR (ATLID), Multi Spectral Imager (MSI) and Broad Band Radiometer (BBR) in addition to Cloud Profiling Radar (CPR). CPR is developed under cooperation of Japanese Aerospace Exploration Agency (JAXA) and National Institute of Information and Communications Technology (NICT) in Japan. The requirement of sensitivity is -35dBZ, therefore CPR uses W-band frequency and needs a large (2.5m) antenna reflector. The large antenna has small footprint and is to give up antenna scanning. From this, some difficulty of external calibration using active radar calibrator (ARC) is recognized. One solution of external calibration is using scattering from natural distributed target, such as sea surface. Then the measurement of sea surface scattering using airborne cloud radar was performed. The sea surface scattering property is being prepared. Second solution is that ARC puts on exact location of sub-satellite track. Precise sub-satellite track prediction is necessary. We focus second solution in this paper. The test experiment was demonstrated using CloudSat of NASA/JPL, which is provided CPR using W-band frequency. The feasibility of this calibration method is discussed.

  19. Optimal calibration method for water distribution water quality model.

    PubMed

    Wu, Zheng Yi

    2006-01-01

    A water quality model is to predict water quality transport and fate throughout a water distribution system. The model is not only a promising alternative for analyzing disinfectant residuals in a cost-effective manner, but also a means of providing enormous engineering insights into the characteristics of water quality variation and constituent reactions. However, a water quality model is a reliable tool only if it predicts what a real system behaves. This paper presents a methodology that enables a modeler to efficiently calibrate a water quality model such that the field observed water quality values match with the model simulated values. The method is formulated to adjust the global water quality parameters and also the element-dependent water quality reaction rates for pipelines and tank storages. A genetic algorithm is applied to optimize the model parameters by minimizing the difference between the model-predicted values and the field-observed values. It is seamlessly integrated with a well-developed hydraulic and water quality modeling system. The approach has provided a generic tool and methodology for engineers to construct the sound water quality model in expedient manner. The method is applied to a real water system and demonstrated that a water quality model can be optimized for managing adequate water supply to public communities. PMID:16854809

  20. Multi-Window Classical Least Squares Multivariate Calibration Methods for Quantitative ICP-AES Analyses

    SciTech Connect

    CHAMBERS,WILLIAM B.; HAALAND,DAVID M.; KEENAN,MICHAEL R.; MELGAARD,DAVID K.

    1999-10-01

    The advent of inductively coupled plasma-atomic emission spectrometers (ICP-AES) equipped with charge-coupled-device (CCD) detector arrays allows the application of multivariate calibration methods to the quantitative analysis of spectral data. We have applied classical least squares (CLS) methods to the analysis of a variety of samples containing up to 12 elements plus an internal standard. The elements included in the calibration models were Ag, Al, As, Au, Cd, Cr, Cu, Fe, Ni, Pb, Pd, and Se. By performing the CLS analysis separately in each of 46 spectral windows and by pooling the CLS concentration results for each element in all windows in a statistically efficient manner, we have been able to significantly improve the accuracy and precision of the ICP-AES analyses relative to the univariate and single-window multivariate methods supplied with the spectrometer. This new multi-window CLS (MWCLS) approach simplifies the analyses by providing a single concentration determination for each element from all spectral windows. Thus, the analyst does not have to perform the tedious task of reviewing the results from each window in an attempt to decide the correct value among discrepant analyses in one or more windows for each element. Furthermore, it is not necessary to construct a spectral correction model for each window prior to calibration and analysis: When one or more interfering elements was present, the new MWCLS method was able to reduce prediction errors for a selected analyte by more than 2 orders of magnitude compared to the worst case single-window multivariate and univariate predictions. The MWCLS detection limits in the presence of multiple interferences are 15 rig/g (i.e., 15 ppb) or better for each element. In addition, errors with the new method are only slightly inflated when only a single target element is included in the calibration (i.e., knowledge of all other elements is excluded during calibration). The MWCLS method is found to be vastly

  1. Improved camera calibration method based on perpendicularity compensation for binocular stereo vision measurement system.

    PubMed

    Jia, Zhenyuan; Yang, Jinghao; Liu, Wei; Wang, Fuji; Liu, Yang; Wang, Lingli; Fan, Chaonan; Zhao, Kai

    2015-06-15

    High-precision calibration of binocular vision systems plays an important role in accurate dimensional measurements. In this paper, an improved camera calibration method is proposed. First, an accurate intrinsic parameters calibration method based on active vision with perpendicularity compensation is developed. Compared to the previous work, this method eliminates the effect of non-perpendicularity of the camera motion on calibration accuracy. The principal point, scale factors, and distortion factors are calculated independently in this method, thereby allowing the strong coupling of these parameters to be eliminated. Second, an accurate global optimization method with only 5 images is presented. The results of calibration experiments show that the accuracy of the calibration method can reach 99.91%. PMID:26193503

  2. Performance comparison of accelerometer calibration algorithms based on 3D-ellipsoid fitting methods.

    PubMed

    Gietzelt, Matthias; Wolf, Klaus-Hendrik; Marschollek, Michael; Haux, Reinhold

    2013-07-01

    Calibration of accelerometers can be reduced to 3D-ellipsoid fitting problems. Changing extrinsic factors like temperature, pressure or humidity, as well as intrinsic factors like the battery status, demand to calibrate the measurements permanently. Thus, there is a need for fast calibration algorithms, e.g. for online analyses. The primary aim of this paper is to propose a non-iterative calibration algorithm for accelerometers with the focus on minimal execution time and low memory consumption. The secondary aim is to benchmark existing calibration algorithms based on 3D-ellipsoid fitting methods. We compared the algorithms regarding the calibration quality and the execution time as well as the number of quasi-static measurements needed for a stable calibration. As evaluation criterion for the calibration, both the norm of calibrated real-life measurements during inactivity and simulation data was used. The algorithms showed a high calibration quality, but the execution time differed significantly. The calibration method proposed in this paper showed the shortest execution time and a very good performance regarding the number of measurements needed to produce stable results. Furthermore, this algorithm was successfully implemented on a sensor node and calibrates the measured data on-the-fly while continuously storing the measured data to a microSD-card. PMID:23566707

  3. A Novel Gaze Tracking Method Based on the Generation of Virtual Calibration Points

    PubMed Central

    Lee, Ji Woo; Heo, Hwan; Park, Kang Ryoung

    2013-01-01

    Most conventional gaze-tracking systems require that users look at many points during the initial calibration stage, which is inconvenient for them. To avoid this requirement, we propose a new gaze-tracking method with four important characteristics. First, our gaze-tracking system uses a large screen located at a distance from the user, who wears a lightweight device. Second, our system requires that users look at only four calibration points during the initial calibration stage, during which four pupil centers are noted. Third, five additional points (virtual pupil centers) are generated with a multilayer perceptron using the four actual points (detected pupil centers) as inputs. Fourth, when a user gazes at a large screen, the shape defined by the positions of the four pupil centers is a distorted quadrangle because of the nonlinear movement of the human eyeball. The gaze-detection accuracy is reduced if we map the pupil movement area onto the screen area using a single transform function. We overcame this problem by calculating the gaze position based on multi-geometric transforms using the five virtual points and the four actual points. Experiment results show that the accuracy of the proposed method is better than that of other methods. PMID:23959241

  4. Real Time Calibration Method for Signal Conditioning Amplifiers

    NASA Technical Reports Server (NTRS)

    Medelius, Pedro J. (Inventor); Mata, Carlos T. (Inventor); Eckhoff, Anthony (Inventor); Perotti, Jose (Inventor); Lucena, Angel (Inventor)

    2004-01-01

    A signal conditioning amplifier receives an input signal from an input such as a transducer. The signal is amplified and processed through an analog to digital converter and sent to a processor. The processor estimates the input signal provided by the transducer to the amplifier via a multiplexer. The estimated input signal is provided as a calibration voltage to the amplifier immediately following the receipt of the amplified input signal. The calibration voltage is amplified by the amplifier and provided to the processor as an amplified calibration voltage. The amplified calibration voltage is compared to the amplified input signal, and if a significant error exists, the gain and/or offset of the amplifier may be adjusted as necessary.

  5. Two-Step Camera Calibration Method Developed for Micro UAV'S

    NASA Astrophysics Data System (ADS)

    Gašparović, M.; Gajski, D.

    2016-06-01

    The development of unmanned aerial vehicles (UAVs) and continuous price reduction of unmanned systems attracted us to this research. Professional measuring systems are dozens of times more expensive and often heavier than "amateur", non-metric UAVs. For this reason, we tested the DJI Phantom 2 Vision Plus UAV. Phantom's smaller mass and velocity can develop less kinetic energy in relation to the professional measurement platforms, which makes it potentially less dangerous for use in populated areas. In this research, we wanted to investigate the ability of such non-metric UAV and find the procedures under which this kind of UAV may be used for the photogrammetric survey. It is important to emphasize that UAV is equipped with an ultra wide-angle camera with 14MP sensor. Calibration of such cameras is a complex process. In the research, a new two-step process is presented and developed, and the results are compared with standard one-step camera calibration procedure. Two-step process involves initially removed distortion on all images, and then uses these images in the phototriangulation with self-calibration. The paper presents statistical indicators which proved that the proposed two-step process is better and more accurate procedure for calibrating those types of cameras than standard one-step calibration. Also, we suggest two-step calibration process as the standard for ultra-wideangle cameras for unmanned aircraft.

  6. A simple and flexible calibration method of non-overlapping camera rig

    NASA Astrophysics Data System (ADS)

    Guan, Banglei; Shang, Yang; Yu, Qifeng; Lei, Zhihui; Zhang, Xiaohu

    2015-05-01

    A simple and flexible method for non-overlapping camera rig calibration that includes camera calibration and relative poses calibration is presented. The proposed algorithm gives the solutions of the cameras parameters and the relative poses simultaneously by using nonlinear optimization. Firstly, the intrinsic and extrinsic parameters of each camera in the rig are estimated individually. Then, a linear solution derived from hand-eye calibration scheme is proposed to compute an initial estimate of the relative poses inside the camera rig. Finally, combined non-linear refinement of all parameters is performed, which optimizes the intrinsic parameters, the extrinsic parameters and relative poses of the coupled camera at the same time. We develop and test a novel approach for calibrating the parameters of non-overlapping camera rig using camera calibration and hand-eye calibration method. The method is designed inter alia for the purpose of deformation measurement using the calibrated rig. Compared the camera calibration with hand-eye calibration separately, our joint calibration is more convenient in practice application. Experimental data shows our algorithm is feasible and effective.

  7. An additional study and implementation of tone calibrated technique of modulation

    NASA Technical Reports Server (NTRS)

    Rafferty, W.; Bechtel, L. K.; Lay, N. E.

    1985-01-01

    The Tone Calibrated Technique (TCT) was shown to be theoretically free from an error floor, and is only limited, in practice, by implementation constraints. The concept of the TCT transmission scheme along with a baseband implementation of a suitable demodulator is introduced. Two techniques for the generation of the TCT signal are considered: a Manchester source encoding scheme (MTCT) and a subcarrier based technique (STCT). The results are summarized for the TCT link computer simulation. The hardware implementation of the MTCT system is addressed and the digital signal processing design considerations involved in satisfying the modulator/demodulator requirements are outlined. The program findings are discussed and future direction are suggested based on conclusions made regarding the suitability of the TCT system for the transmission channel presently under consideration.

  8. Optical geometry calibration method for free-form digital tomosynthesis

    NASA Astrophysics Data System (ADS)

    Chtcheprov, Pavel; Hartman, Allison; Shan, Jing; Lee, Yueh Z.; Zhou, Otto; Lu, Jianping

    2016-03-01

    Digital tomosynthesis is a type of limited angle tomography that allows 3D information to be reconstructed from a set of x-ray projection images taken at various angles using an x-ray tube, a mechanical arm to rotate the tube about the object, and a digital detector. Tomosynthesis reconstruction requires the precise location of the detector with respect to each x-ray source, forcing all current clinical tomosynthesis systems to use a physically coupled source and detector so the geometry is always known and is always the same. This limits the imaging geometries and its large size is impractical for mobile or field operations. To counter this, we have developed a free form tomosynthesis with a decoupled, free-moving source and detector that uses a novel optical method for accurate and real-time geometry calibration to allow for manual, hand-held tomosynthesis and even CT imaging. We accomplish this by using a camera, attached to the source, to track the motion of the source relative to the detector. Attached to the detector is an optical pattern and the image captured by the camera is then used to determine the relative camera/pattern position and orientation by analyzing the pattern distortion and calculating the source positions for each projection, necessary for 3D reconstruction. This allows for portable imaging in the field and also as an inexpensive upgrade to existing 2D systems, such as in developing countries, to provide 3D image data. Here we report the first feasibility demonstrations of free form digital tomosynthesis systems using the method.

  9. Research on method of geometry and spectral calibration of pushbroom dispersive hyperspectral imager

    NASA Astrophysics Data System (ADS)

    He, Zhiping; Shu, Rong; Wang, Jianyu

    2012-11-01

    Development and application of airborne and aerospace hyperspectral imager press for high precision geometry and spectral calibration of pixels of image cube. The research of geometry and spectral calibration of pushbroom hyperspectral imager, its target is giving the coordinate of angle field of view and center wavelength of each detect unit in focal plane detector of hyperspectral imager, and achieves the high precision, full field of view, full channel geometry and spectral calibration. It is importance for imaging quantitative and deep application of hyperspectal imager. The paper takes the geometry and spectral calibration of pushbroom dispersive hyperspectral imager as case study, and research on the constitution and analysis of imaging mathematical model. Aimed especially at grating-dispersive hyperspectral imaging, the specialty of the imaging mode and dispersive method has been concretely analyzed. Based on the analysis, the theory and feasible method of geometry and spectral calibration of dispersive hyperspectral imager is set up. The key technique has been solved is As follows: 1). the imaging mathematical model and feasible method of geometry and spectral calibration for full pixels of image cube has been set up, the feasibility of the calibration method has been analyzed. 2). the engineering model and method of the geometry and spectral calibration of pushbroom dispersive hyperspectral imager has been set up and the calibration equipment has been constructed, and the calibration precision has been analyzed.

  10. Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories.

    PubMed

    Hourdakis, Costas J; Büermann, Ludwig; Ciraj-Bjelac, Olivera; Csete, Istvan; Delis, Harry; Gomola, Igor; Persson, Linda; Novak, Leos; Petkov, Ivailo; Toroi, Paula

    2016-01-01

    A comparison of calibration results and procedures in terms of air kerma length product, PKL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the "partial irradiation method" is the preferred method for the pencil-type IC calibration in terms of PKL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement (>99%) for both dosimetry quantities (PKL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed. PMID:26508013

  11. Model Robust Calibration: Method and Application to Electronically-Scanned Pressure Transducers

    NASA Technical Reports Server (NTRS)

    Walker, Eric L.; Starnes, B. Alden; Birch, Jeffery B.; Mays, James E.

    2010-01-01

    This article presents the application of a recently developed statistical regression method to the controlled instrument calibration problem. The statistical method of Model Robust Regression (MRR), developed by Mays, Birch, and Starnes, is shown to improve instrument calibration by reducing the reliance of the calibration on a predetermined parametric (e.g. polynomial, exponential, logarithmic) model. This is accomplished by allowing fits from the predetermined parametric model to be augmented by a certain portion of a fit to the residuals from the initial regression using a nonparametric (locally parametric) regression technique. The method is demonstrated for the absolute scale calibration of silicon-based pressure transducers.

  12. Evaluating Three Different Methods of Determining Addition in Presbyopia

    PubMed Central

    Yazdani, Negareh; Khorasani, Abbas Azimi; Moghadam, Hanieh Mirhajian; Yekta, Abbas Ali; Ostadimoghaddam, Hadi; Shandiz, Javad Heravian

    2016-01-01

    Purpose: To compare three different methods for determining addition in presbyopes. Methods: The study included 81 subjects with presbyopia who aged 40-70 years. Reading addition values were measured using 3 approaches including the amplitude of accommodation (AA), dynamic retinoscopy (DR), and increasing plus lens (IPL). Results: IPL overestimated reading addition relative to other methods. Mean near addition obtained by AA, DR and IPL were 1.31, 1.68 and 1.77, respectively. Our results showed that IPL method could provide 20/20 vision at near in the majority of presbyopic subjects (63.4%). Conclusion: The results were approximately the same for 3 methods and provided comparable final addition; however, mean near additions were higher with increasing plus lens compared with the other two methods. In presbyopic individuals, increasing plus lens is recommended as the least time-consuming method with the range of ±0.50 diopter at the 40 cm working distance. PMID:27621785

  13. Calibration of AVHRR sensors using the reflectance-based method

    NASA Astrophysics Data System (ADS)

    Czapla-Myers, Jeffrey S.; Thome, Kurtis J.; Leisso, Nathan P.

    2007-09-01

    The Remote Sensing Group at the University of Arizona has been active in the vicarious calibration of numerous sensors through the use of ground-based test sites. Recent efforts have included work to develop cross-calibration information between these sensors using the results from the reflectance-based approach. The current work extends the cross-calibration to the AVHRR series of sensors, specifically NOAA-17, and NOAA-18. The results include work done based on data collected by ground-based personnel nearly coincident with the sensor overpasses. The available number of calibrations for the AVHRR series is increased through a set of ground-based radiometers that are deployed without the need for on-site personnel and have been operating for more than three years at Railroad Valley Playa. The spectral, spatial, and temporal characteristics of the 1-km2 large-footprint site at Railroad Valley are well understood. It is therefore well suited for the radiometric calibration of AVHRR, which has a nadir-viewing footprint of 1.1 x 1.1 km. The at-sensor radiance is predicted via a radiative transfer code using atmospheric data from a fully-automated solar radiometer. The results for AVHRR show that errors are currently larger for the automated data sets, but results indicate that the AVHRR sensors studied in this work are consistent with the Aqua and Terra MODIS sensors to within the uncertainties of each sensor.

  14. Indoor Calibration for Stereoscopic Camera STC, A New Method

    NASA Astrophysics Data System (ADS)

    Simioni, E.; Re, C.; Da Deppo, V.; Naletto, G.; Borrelli, D.; Dami, M.; Ficai Veltroni, I.; Cremonese, G.

    2014-10-01

    In the framework of the ESA-JAXA BepiColombo mission to Mercury, the global mapping of the planet will be performed by the on-board Stereo Camera (STC), part of the SIMBIO-SYS suite [1]. In this paper we propose a new technique for the validation of the 3D reconstruction of planetary surface from images acquired with a stereo camera. STC will provide a three-dimensional reconstruction of Mercury surface. The generation of a DTM of the observed features is based on the processing of the acquired images and on the knowledge of the intrinsic and extrinsic parameters of the optical system. The new stereo concept developed for STC needs a pre-flight verification of the actual capabilities to obtain elevation information from stereo couples: for this, a stereo validation setup to get an indoor reproduction of the flight observing condition of the instrument would give a much greater confidence to the developed instrument design. STC is the first stereo satellite camera with two optical channels converging in a unique sensor. Its optical model is based on a brand new concept to minimize mass and volume and to allow push-frame imaging. This model imposed to define a new calibration pipeline to test the reconstruction method in a controlled ambient. An ad-hoc indoor set-up has been realized for validating the instrument designed to operate in deep space, i.e. in-flight STC will have to deal with source/target essentially placed at infinity. This auxiliary indoor setup permits on one side to rescale the stereo reconstruction problem from the operative distance in-flight of 400 km to almost 1 meter in lab; on the other side it allows to replicate different viewing angles for the considered targets. Neglecting for sake of simplicity the Mercury curvature, the STC observing geometry of the same portion of the planet surface at periherm corresponds to a rotation of the spacecraft (SC) around the observed target by twice the 20° separation of each channel with respect to nadir

  15. Absolute radiometric calibration of the RapidEye multispectral imager using the reflectance-based vicarious calibration method

    NASA Astrophysics Data System (ADS)

    Naughton, Denis; Brunn, Andreas; Czapla-Myers, Jeff; Douglass, Scott; Thiele, Michael; Weichelt, Horst; Oxfort, Michael

    2011-01-01

    RapidEye AG is a commercial provider of geospatial information products and customized solutions derived from Earth observation image data. The source of the data is the RapidEye constellation consisting of five low-earth-orbit imaging satellites. We describe the rationale, methods, and results of a reflectance-based vicarious calibration campaign that was conducted between April 2009 and May 2010 at Railroad Valley Playa and Ivanpah Playa to determine the on-orbit radiometric accuracy of the RapidEye sensor. In situ surface spectral reflectance measurements of known ground targets and an assessment of the atmospheric conditions above the sites were taken during spacecraft overpasses. The ground data are used as input to a radiative transfer code to compute a band-specific top-of-atmosphere spectral radiance. A comparison of these predicted values based on absolute physical data to the measured at-sensor spectral radiance provide the absolute calibration of the sensor. Initial assessments show that the RapidEye sensor response is within 8% of the predicted values. Outcomes from this campaign are then used to update the calibration parameters in the ground segment processing system. Subsequent verification events confirmed that the measured RapidEye response improved to within 4% of the predictions based on the vicarious calibration method.

  16. 10 CFR 2.705 - Discovery-additional methods.

    Code of Federal Regulations, 2011 CFR

    2011-01-01

    ... 10 Energy 1 2011-01-01 2011-01-01 false Discovery-additional methods. 2.705 Section 2.705 Energy NUCLEAR REGULATORY COMMISSION RULES OF PRACTICE FOR DOMESTIC LICENSING PROCEEDINGS AND ISSUANCE OF ORDERS Rules for Formal Adjudications § 2.705 Discovery-additional methods. (a) Discovery methods. Parties may obtain discovery by one or more of...

  17. Calibration of three rainfall simulators with automatic measurement methods

    NASA Astrophysics Data System (ADS)

    Roldan, Margarita

    2010-05-01

    CALIBRATION OF THREE RAINFALL SIMULATORS WITH AUTOMATIC MEASUREMENT METHODS M. Roldán (1), I. Martín (2), F. Martín (2), S. de Alba(3), M. Alcázar(3), F.I. Cermeño(3) 1 Grupo de Investigación Ecología y Gestión Forestal Sostenible. ECOGESFOR-Universidad Politécnica de Madrid. E.U.I.T. Forestal. Avda. Ramiro de Maeztu s/n. Ciudad Universitaria. 28040 Madrid. margarita.roldan@upm.es 2 E.U.I.T. Forestal. Avda. Ramiro de Maeztu s/n. Ciudad Universitaria. 28040 Madrid. 3 Facultad de Ciencias Geológicas. Universidad Complutense de Madrid. Ciudad Universitaria s/n. 28040 Madrid The rainfall erosivity is the potential ability of rain to cause erosion. It is function of the physical characteristics of rainfall (Hudson, 1971). Most expressions describing erosivity are related to kinetic energy or momentum and so with drop mass or size and fall velocity. Therefore, research on factors determining erosivity leds to the necessity to study the relation between fall height and fall velocity for different drop sizes, generated in a rainfall simulator (Epema G.F.and Riezebos H.Th, 1983) Rainfall simulators are one of the most used tools for erosion studies and are used to determine fall velocity and drop size. Rainfall simulators allow repeated and multiple measurements The main reason for use of rainfall simulation as a research tool is to reproduce in a controlled way the behaviour expected in the natural environment. But in many occasions when simulated rain is used in order to compare it with natural rain, there is a lack of correspondence between natural and simulated rain and this can introduce some doubt about validity of data because the characteristics of natural rain are not adequately represented in rainfall simulation research (Dunkerley D., 2008). Many times the rainfall simulations have high rain rates and they do not resemble natural rain events and these measures are not comparables. And besides the intensity is related to the kinetic energy which

  18. 40 CFR 80.157 - Volumetric additive reconciliation (“VAR”), equipment calibration, and recordkeeping requirements.

    Code of Federal Regulations, 2012 CFR

    2012-07-01

    ... limited number of specified people those who have the ability to alter or delete the records. In addition... transfers from the facility of unadditized base gasoline; (3) For all detergent blending facilities,...

  19. Methods to calibrate the absolute receive sensitivity of single-element, focused transducers

    PubMed Central

    Rich, Kyle T.; Mast, T. Douglas

    2015-01-01

    Absolute pressure measurements of acoustic emissions by single-element, focused passive cavitation detectors would be facilitated by improved wideband receive calibration techniques. Here, calibration methods were developed to characterize the absolute, frequency-dependent receive sensitivity of a spherically focused, single-element transducer using pulse-echo and pitch-catch techniques. Validation of these calibration methods on a focused receiver were made by generating a pulse from a small diameter source at the focus of the transducer and comparing the absolute pressure measured by a calibrated hydrophone to that of the focused transducer using the receive sensitivities determined here. PMID:26428812

  20. Methods to calibrate the absolute receive sensitivity of single-element, focused transducers.

    PubMed

    Rich, Kyle T; Mast, T Douglas

    2015-09-01

    Absolute pressure measurements of acoustic emissions by single-element, focused passive cavitation detectors would be facilitated by improved wideband receive calibration techniques. Here, calibration methods were developed to characterize the absolute, frequency-dependent receive sensitivity of a spherically focused, single-element transducer using pulse-echo and pitch-catch techniques. Validation of these calibration methods on a focused receiver were made by generating a pulse from a small diameter source at the focus of the transducer and comparing the absolute pressure measured by a calibrated hydrophone to that of the focused transducer using the receive sensitivities determined here. PMID:26428812

  1. A method for intensity calibration of an electron spectrometer with multi-angle detection

    NASA Astrophysics Data System (ADS)

    Ábrók, Levente; Buhr, Ticia; Kövér, Ákos; Balog, Róbert; Hatvani, Dávid; Herczku, Péter; Kovács, Sándor; Ricz, Sándor

    2016-02-01

    A special electrostatic electron spectrometer designed for precise and unique experiments and an intensity calibration method for universal application in electron spectroscopy are presented. The upgrade of the analyzer enables the intensity calibration at arbitrary electron energies using elastically scattered electrons. In order to test the calibration procedure the double differential (in energy and angle) ionization cross sections (DDCS) of electrons ejected from 300 keV proton-argon collisions were measured and compared with the data of Rudd et al. [1]. The good agreement between the two data sets verifies the applicability of the calibration method.

  2. Calibration method for division of focal plane polarimeters in the optical and near-infrared regime

    NASA Astrophysics Data System (ADS)

    York, Timothy; Gruev, Viktor

    2011-06-01

    Advances in nanofabrication allow for the creation of metallic nanowires acting as linear polarizers in the visible and near infrared regime. The monolithic integration of silicon detectors and pixelated nanowire metallic polarization filters allows for an efficient realization of high resolution polarization imaging sensors. These silicon sensors, known as division of focal plane polarimeters, capture polarization information of the imaged environment from ~400nm to 1050nm wavelength. The performance of the polarization sensor can be degraded by both irregularities in the fabrication of the nanowires and possible misalignment errors during the final deposition of the optical nanowire filters on the surface of the imaging sensor. In addition, electronic offsets due to the readout circuitry, electronic crosstalk, and optical crosstalk will also negatively affect the quality of the polarization information. Partial compensation for many of these post-fabrication errors can be accomplished through the use of a camera calibration routine. This paper will describe one such routine, and show how its application can increase the quality of measurements in both the degree of linear polarization and angle of polarization in the visible spectrum. The imaging array of the division of focal plane polarimeter is segmented into two by two blocks of superpixels. The calibration method chooses one of the four pixels as a reference, and then a gain and offset for each of the remaining three is computed based on this reference. The output is a calibration matrix for each pixel in the image array.

  3. Absolute calibration of OH density in a nanosecond pulsed plasma filament in atmospheric pressure He-H2O: comparison of independent calibration methods

    NASA Astrophysics Data System (ADS)

    Verreycken, T.; van der Horst, R. M.; Sadeghi, N.; Bruggeman, P. J.

    2013-11-01

    The absolute density of OH radicals generated in a nanosecond pulsed filamentary discharge in atmospheric pressure He +0.84% H2O is measured independently by UV absorption and laser induced fluorescence (LIF) calibrated with Rayleigh scattering. For the calibration of LIF with Rayleigh scattering, two LIF models, with six levels and four levels, are studied to investigate the influence of the rotational and vibrational energy transfers. In addition, a chemical model is used to deduce the OH density in the afterglow from the relative LIF intensity as function of time. The different models show good correspondence and by comparing these different methods, the accuracy and the effect of assumptions on the obtained OH density are discussed in detail. This analysis includes an analysis of the sensitivity to parameters used in the LIF models.

  4. An optimization-based method for geometrical calibration in cone-beam CT without dedicated phantoms

    NASA Astrophysics Data System (ADS)

    Panetta, D.; Belcari, N.; DelGuerra, A.; Moehrs, S.

    2008-07-01

    In this paper we present a new method for the determination of geometrical misalignments in cone-beam CT scanners, from the analysis of the projection data of a generic object. No a priori knowledge of the object shape and positioning is required. We show that a cost function, which depends on the misalignment parameters, can be defined using the projection data and that such a cost function has a local minimum in correspondence to the actual parameters of the system. Hence, the calibration of the scanner can be carried out by minimizing the cost function using standard optimization techniques. The method is developed for a particular class of 3D object functions, for which the redundancy of the fan beam sinogram in the transaxial midplane can be extended to cone-beam projection data, even at wide cone angles. The method has an approximated validity for objects which do not belong to that class; in that case, a suitable subset of the projection data can be selected in order to compute the cost function. We show by numerical simulations that our method is capable to determine with high accuracy the most critical misalignment parameters of the scanner, i.e., the transversal shift and the skew of the detector. Additionally, the detector slant can be determined. Other parameters such as the detector tilt, the longitudinal shift and the error in the source-detector distance cannot be determined with our method, as the proposed cost function has a very weak dependence on them. However, due to the negligible influence of these latter parameters in the reconstructed image quality, they can be kept fixed at estimated values in both calibration and reconstruction processes without compromising the final result. A trade-off between computational cost and calibration accuracy must be considered when choosing the data subset used for the computation of the cost function. Results on real data of a mouse femur as obtained with a small animal micro-CT are shown as well, proving

  5. Method and apparatus for calibrating a tiled display

    NASA Technical Reports Server (NTRS)

    Johnson, Michael J. (Inventor); Chen, Chung-Jen (Inventor); Chandrasekhar, Rajesh (Inventor)

    2001-01-01

    A display system that can be calibrated and re-calibrated with a minimal amount of manual intervention. To accomplish this, one or more cameras are provided to capture an image of the display screen. The resulting captured image is processed to identify any non-desirable characteristics, including visible artifacts such as seams, bands, rings, etc. Once the non-desirable characteristics are identified, an appropriate transformation function is determined. The transformation function is used to pre-warp the input video signal that is provided to the display such that the non-desirable characteristics are reduced or eliminated from the display. The transformation function preferably compensates for spatial non-uniformity, color non-uniformity, luminance non-uniformity, and other visible artifacts.

  6. Method and Apparatus for Accurately Calibrating a Spectrometer

    NASA Technical Reports Server (NTRS)

    Youngquist, Robert C. (Inventor); Simmons, Stephen M. (Inventor)

    2013-01-01

    A calibration assembly for a spectrometer is provided. The assembly includes a spectrometer having n detector elements, where each detector element is assigned a predetermined wavelength value. A first source emitting first radiation is used to calibrate the spectrometer. A device is placed in the path of the first radiation to split the first radiation into a first beam and a second beam. The assembly is configured so that one of the first and second beams travels a path-difference distance longer than the other of the first and second beams. An output signal is generated by the spectrometer when the first and second beams enter the spectrometer. The assembly includes a controller operable for processing the output signal and adapted to calculate correction factors for the respective predetermined wavelength values assigned to each detector element.

  7. A New Method for Common Calibration of Sun-Sky-Lunar Photometer

    NASA Astrophysics Data System (ADS)

    Li, Kaitao; Li, Zhengqiang; Li, Donghui; Xu, Hua; Xie, Yisong; Li, Li; Chen, Xingfeng; Ma, Yan

    2016-04-01

    A new calibration method is introduced to transfer extraterrestrial calibration coefficients to the moon measurements for a new sun-sky-lunar photometer, trade name CE318-T from CIMEL. The new transfer method has no relationship with lunar phase, therefore, the precision of the results is improved, and error analysis suggests that the uncertainty of the transferred method is about 2.2-2.6%, smaller than the lunar Langley calibrations. At the same time, the calibration time is also saved. The Sun-Sky-Lunar photometer numbered #1202 and located on the roof of Institute of remote sensing and digital earth (RADI) in Beijing was used in this study. The extraterrestrial calibration coefficients were got by using Langley calibration performed at Ali with a height of 5053 m above sea level in Tibet. The new lunar calibration coefficients were obtained with the new transfer method. And then the nocturnal AODs were calculated, which are well consistent with the daytime observations. The differences between two AODs obtained with transferred calibration coefficients and lunar Langley method were also compared in this paper. In this study, Lidar observation results was also presented to compare with the lunar observations, the results show that the nocturnal AODs have the same variation tendency with the Lidar observations.

  8. A long-baseline method for HST gyro drift rate bias calibration

    NASA Technical Reports Server (NTRS)

    Boia, John J.; Welter, Gary L.; Gakenheimer, Martin F.

    1994-01-01

    The routine on-orbit calibration of the Hubble Space Telescope (HST) rate gyro assemblies (RGA's) has depended on several related algorithms for drift rate bias calibration. The gyros have exhibited time-varying biases, which must be regularly corrected to maintain pointing stability. Currently, gyro drift parameters are uplinked to the spacecraft every 1-2 days for low rate mode and every 7 days for high rate mode. In order to minimize the impact of frequent calibrations on the HST science schedule, we have refined the gyro calibration algorithms and data collection schemes to reduce the amounts of telemetry data and processing time required for accurate bias calibration. We present a review of the evolution of the gyro calibration algorithms, with particular attention to what we have called the long-baseline bias (LBBIAS) technique, and describe the relative success of these methods in maintaining spacecraft stability.

  9. Camera calibration method of binocular stereo vision based on OpenCV

    NASA Astrophysics Data System (ADS)

    Zhong, Wanzhen; Dong, Xiaona

    2015-10-01

    Camera calibration, an important part of the binocular stereo vision research, is the essential foundation of 3D reconstruction of the spatial object. In this paper, the camera calibration method based on OpenCV (open source computer vision library) is submitted to make the process better as a result of obtaining higher precision and efficiency. First, the camera model in OpenCV and an algorithm of camera calibration are presented, especially considering the influence of camera lens radial distortion and decentering distortion. Then, camera calibration procedure is designed to compute those parameters of camera and calculate calibration errors. High-accurate profile extraction algorithm and a checkboard with 48 corners have also been used in this part. Finally, results of calibration program are presented, demonstrating the high efficiency and accuracy of the proposed approach. The results can reach the requirement of robot binocular stereo vision.

  10. An automated calibration method for non-see-through head mounted displays.

    PubMed

    Gilson, Stuart J; Fitzgibbon, Andrew W; Glennerster, Andrew

    2011-08-15

    Accurate calibration of a head mounted display (HMD) is essential both for research on the visual system and for realistic interaction with virtual objects. Yet, existing calibration methods are time consuming and depend on human judgements, making them error prone, and are often limited to optical see-through HMDs. Building on our existing approach to HMD calibration Gilson et al. (2008), we show here how it is possible to calibrate a non-see-through HMD. A camera is placed inside a HMD displaying an image of a regular grid, which is captured by the camera. The HMD is then removed and the camera, which remains fixed in position, is used to capture images of a tracked calibration object in multiple positions. The centroids of the markers on the calibration object are recovered and their locations re-expressed in relation to the HMD grid. This allows established camera calibration techniques to be used to recover estimates of the HMD display's intrinsic parameters (width, height, focal length) and extrinsic parameters (optic centre and orientation of the principal ray). We calibrated a HMD in this manner and report the magnitude of the errors between real image features and reprojected features. Our calibration method produces low reprojection errors without the need for error-prone human judgements. PMID:21620891

  11. Calibration to surface reflectance of terrestrial imaging spectrometry data: Comparison of methods

    NASA Technical Reports Server (NTRS)

    Clark, Roger N.; Swayze, Gregg A.; Heidebrecht, Kathy; Green, Robert O.; Goetz, F. H.

    1995-01-01

    Many algorithms for spectral analysis of imaging spectroscopy data of the Earth's surface require that the data be calibrated to surface reflectance. Calibration requires removing instrumental response, solar irradiance, atmospheric transmittance, and atmospheric scattering from the radiance detected at the sensor. Depending on the amount of support data, this can be a formidable task. This paper examines four methods of calibration: (1) a radiative transfer model from the University of Colorado (ATREM: Gao and Goetz, 1990; Gao et al., 1992), (2) a MODTRAN-based method developed at the Jet Propulsion Lab by Green et al., (1191), (3) a ground calibration using known sites as standards, and (4) a combined approach using radiative transfer methods and ground calibration. Data from the Airborne Visual and Infra-Red Imaging Spectrometer (AVIRIS) instrument were evaluated from data sets obtained over multiple years and multiple sites.

  12. Improvement of Accuracy in Environmental Dosimetry by TLD Cards Using Three-dimensional Calibration Method

    PubMed Central

    HosseiniAliabadi, S. J.; Hosseini Pooya, S. M.; Afarideh, H.; Mianji, F.

    2015-01-01

    Introduction The angular dependency of response for TLD cards may cause deviation from its true value on the results of environmental dosimetry, since TLDs may be exposed to radiation at different angles of incidence from the surrounding area. Objective A 3D setting of TLD cards has been calibrated isotropically in a standard radiation field to evaluate the improvement of the accuracy of measurement for environmental dosimetry. Method Three personal TLD cards were rectangularly placed in a cylindrical holder, and calibrated using 1D and 3D calibration methods. Then, the dosimeter has been used simultaneously with a reference instrument in a real radiation field measuring the accumulated dose within a time interval. Result The results show that the accuracy of measurement has been improved by 6.5% using 3D calibration factor in comparison with that of normal 1D calibration method. Conclusion This system can be utilized in large scale environmental monitoring with a higher accuracy. PMID:26157729

  13. A New Method of Theodolite Calibration Based on Image Processing Technology

    NASA Astrophysics Data System (ADS)

    Zou, Hui-Hui; Wu, Hong-Bing; Chen, Di

    Aiming at improving the theodolite calibration method for space tracking ship, a calibration device which consists of hardware and software is designed in this paper. Hereinto, the hardware part is a set of optical acquisition system that includes CCD, lens and 0.2" collimator, while the software part contains image acquisition module, image processing module, data processing module and interface display module. During the calibration process, the new methods of image denoising and image character extraction are applied to improve the precision of image measure. The result of the experiment shows that the calibration criteria of the theodolite errors was met by applying the image processing technology of the theodolite calibration device, it is more accurate than the manual reading method under the same situation in dock.

  14. Fast wavelength calibration method for spectrometers based on waveguide comb optical filter

    SciTech Connect

    Yu, Zhengang; Huang, Meizhen Zou, Ye; Wang, Yang; Sun, Zhenhua; Cao, Zhuangqi

    2015-04-15

    A novel fast wavelength calibration method for spectrometers based on a standard spectrometer and a double metal-cladding waveguide comb optical filter (WCOF) is proposed and demonstrated. By using the WCOF device, a wide-spectrum beam is comb-filtered, which is very suitable for spectrometer wavelength calibration. The influence of waveguide filter’s structural parameters and the beam incident angle on the comb absorption peaks’ wavelength and its bandwidth are also discussed. The verification experiments were carried out in the wavelength range of 200–1100 nm with satisfactory results. Comparing with the traditional wavelength calibration method based on discrete sparse atomic emission or absorption lines, the new method has some advantages: sufficient calibration data, high accuracy, short calibration time, fit for produce process, stability, etc.

  15. Performance analysis for time-frequency MUSIC algorithm in presence of both additive noise and array calibration errors

    NASA Astrophysics Data System (ADS)

    Khodja, Mohamed; Belouchrani, Adel; Abed-Meraim, Karim

    2012-12-01

    This article deals with the application of Spatial Time-Frequency Distribution (STFD) to the direction finding problem using the Multiple Signal Classification (MUSIC)algorithm. A comparative performance analysis is performed for the method under consideration with respect to that using data covariance matrix when the received array signals are subject to calibration errors in a non-stationary environment. An unified analytical expression of the Direction Of Arrival (DOA) error estimation is derived for both methods. Numerical results show the effect of the parameters intervening in the derived expression on the algorithm performance. It is particularly observed that for low Signal to Noise Ratio (SNR) and high Signal to sensor Perturbation Ratio (SPR) the STFD method gives better performance, while for high SNR and for the same SPR both methods give similar performance.

  16. The molecular branching ratio method for calibration of optical systems in the vacuum ultraviolet

    NASA Technical Reports Server (NTRS)

    Mumma, M. J.

    1972-01-01

    The intensity distribution of bands belonging to six molecular band systems is discussed with special emphasis on their usefulness for intensity calibration of optical systems in the vacuum ultraviolet (1000A Lambda 3000A). The theory of molecular band intensities is outlined and the technique of measuring the spectral response curve is described. Several methods for establishing an absolute intensity calibration are discussed.

  17. A Study for Efficient Methods of System Calibration between Optical and Range Sensors

    NASA Astrophysics Data System (ADS)

    Choi, W.; Kim, C.; Kim, Y.

    2015-06-01

    Recently, interests in 3D indoor modeling and positioning have been growing. Data fusion by using different sensors data is one of the 3D model producing methods. For a data fusion between two kinds of sensors, precise system calibration is essential. If relative geometric location of each sensor can be accurately measured with a system-calibration, it is possible to locate a pixel that corresponds to the same object in two different images, and thus, produce a more precise data-fusion. Purpose of this study is finding more efficient method of system calibration between optical and range sensor. For this purpose, experiment was designed by considering following variables, i) system calibration method, ii) testbed type, iii) and distance data(whether use it or not). So, In this study, test-bed for system calibration was designed by considering the characteristics of sensors. Also, precise simulation was done to find efficient method of system calibration, and its results were reflected in real experiment. Results of simulation show that the bundle adjustment method is more efficient than single photo resection in system calibration between range and optical sensors. And the most efficient case was when using i) the bundle adjustment with ii) the simulated data set which were obtained between 2m to 4m away from the test-bed. These results of simulation were reflected in real system calibration. Finally, real system calibration were performed and its results were compared to results of simulation. And accuracy of system calibration was evaluated by producing fusion data between range and optical sensors.

  18. In Search of Easy-to-Use Methods for Calibrating ADCP's for Velocity and Discharge Measurements

    USGS Publications Warehouse

    Oberg, K.

    2002-01-01

    A cost-effective procedure for calibrating acoustic Doppler current profilers (ADCP) in the field was presented. The advantages and disadvantages of various methods which are used for calibrating ADCP were discussed. The proposed method requires the use of differential global positioning system (DGPS) with sub-meter accuracy and standard software for collecting ADCP data. The method involves traversing a long (400-800 meter) course at a constant compass heading and speed, while collecting simultaneous DGPS and ADCP data.

  19. An Accurate Calibration Method Based on Velocity in a Rotational Inertial Navigation System

    PubMed Central

    Zhang, Qian; Wang, Lei; Liu, Zengjun; Feng, Peide

    2015-01-01

    Rotation modulation is an effective method to enhance the accuracy of an inertial navigation system (INS) by modulating the gyroscope drifts and accelerometer bias errors into periodically varying components. The typical RINS drives the inertial measurement unit (IMU) rotation along the vertical axis and the horizontal sensors’ errors are modulated, however, the azimuth angle error is closely related to vertical gyro drift, and the vertical gyro drift also should be modulated effectively. In this paper, a new rotation strategy in a dual-axis rotational INS (RINS) is proposed and the drifts of three gyros could be modulated, respectively. Experimental results from a real dual-axis RINS demonstrate that the maximum azimuth angle error is decreased from 0.04° to less than 0.01° during 1 h. Most importantly, the changing of rotation strategy leads to some additional errors in the velocity which is unacceptable in a high-precision INS. Then the paper studies the basic reason underlying horizontal velocity errors in detail and a relevant new calibration method is designed. Experimental results show that after calibration and compensation, the fluctuation and stages in the velocity curve disappear and velocity precision is improved. PMID:26225983

  20. An Accurate Calibration Method Based on Velocity in a Rotational Inertial Navigation System.

    PubMed

    Zhang, Qian; Wang, Lei; Liu, Zengjun; Feng, Peide

    2015-01-01

    Rotation modulation is an effective method to enhance the accuracy of an inertial navigation system (INS) by modulating the gyroscope drifts and accelerometer bias errors into periodically varying components. The typical RINS drives the inertial measurement unit (IMU) rotation along the vertical axis and the horizontal sensors' errors are modulated, however, the azimuth angle error is closely related to vertical gyro drift, and the vertical gyro drift also should be modulated effectively. In this paper, a new rotation strategy in a dual-axis rotational INS (RINS) is proposed and the drifts of three gyros could be modulated, respectively. Experimental results from a real dual-axis RINS demonstrate that the maximum azimuth angle error is decreased from 0.04° to less than 0.01° during 1 h. Most importantly, the changing of rotation strategy leads to some additional errors in the velocity which is unacceptable in a high-precision INS. Then the paper studies the basic reason underlying horizontal velocity errors in detail and a relevant new calibration method is designed. Experimental results show that after calibration and compensation, the fluctuation and stages in the velocity curve disappear and velocity precision is improved. PMID:26225983

  1. An Improved Calibration Method for Hydrazine Monitors for the United States Air Force

    SciTech Connect

    Korsah, K

    2003-07-07

    This report documents the results of Phase 1 of the ''Air Force Hydrazine Detector Characterization and Calibration Project''. A method for calibrating model MDA 7100 hydrazine detectors in the United States Air Force (AF) inventory has been developed. The calibration system consists of a Kintek 491 reference gas generation system, a humidifier/mixer system which combines the dry reference hydrazine gas with humidified diluent or carrier gas to generate the required humidified reference for calibrations, and a gas sampling interface. The Kintek reference gas generation system itself is periodically calibrated using an ORNL-constructed coulometric titration system to verify the hydrazine concentration of the sample atmosphere in the interface module. The Kintek reference gas is then used to calibrate the hydrazine monitors. Thus, coulometric titration is only used to periodically assess the performance of the Kintek reference gas generation system, and is not required for hydrazine monitor calibrations. One advantage of using coulometric titration for verifying the concentration of the reference gas is that it is a primary standard (if used for simple solutions), thereby guaranteeing, in principle, that measurements will be traceable to SI units (i.e., to the mole). The effect of humidity of the reference gas was characterized by using the results of concentrations determined by coulometric titration to develop a humidity correction graph for the Kintek 491 reference gas generation system. Using this calibration method, calibration uncertainty has been reduced by 50% compared to the current method used to calibrate hydrazine monitors in the Air Force inventory and calibration time has also been reduced by more than 20%. Significant findings from studies documented in this report are the following: (1) The Kintek 491 reference gas generation system (generator, humidifier and interface module) can be used to calibrate hydrazine detectors. (2) The Kintek system output

  2. Self-calibration method for rotating laser positioning system using interscanning technology and ultrasonic ranging.

    PubMed

    Wu, Jun; Yu, Zhijing; Zhuge, Jingchang

    2016-04-01

    A rotating laser positioning system (RLPS) is an efficient measurement method for large-scale metrology. Due to multiple transmitter stations, which consist of a measurement network, the position relationship of these stations must be first calibrated. However, with such auxiliary devices such as a laser tracker, scale bar, and complex calibration process, the traditional calibration methods greatly reduce the measurement efficiency. This paper proposes a self-calibration method for RLPS, which can automatically obtain the position relationship. The method is implemented through interscanning technology by using a calibration bar mounted on the transmitter station. Each bar is composed of three RLPS receivers and one ultrasonic sensor whose coordinates are known in advance. The calibration algorithm is mainly based on multiplane and distance constraints and is introduced in detail through a two-station mathematical model. The repeated experiments demonstrate that the coordinate measurement uncertainty of spatial points by using this method is about 0.1 mm, and the accuracy experiments show that the average coordinate measurement deviation is about 0.3 mm compared with a laser tracker. The accuracy can meet the requirements of most applications, while the calibration efficiency is significantly improved. PMID:27140762

  3. Camera calibration method for dimensional measurement of heavy forging in large scale

    NASA Astrophysics Data System (ADS)

    Liu, Bin; Hu, Chunhai; Song, Xiaoxue; Zhao, Zhenqing

    2009-07-01

    Camera calibration method plays an important role in the stereovision system to resolve the problems of dimensional measurement of heavy forging. Due to the intensive vibrating, the camera parameters must be calibrated every time after the action of the water press. This paper presents a method using the scene geometry to calibrate cameras. In the context of heavy machinery environments, the constraints which can be used are parallelism and orthogonality. These constraints lead to geometrically intuitive methods to calibrate the cameras. The huge forging equipment such as water press belongs to geometrically constrained object and insusceptible to vibrating, which gives natural prior knowledge and constraint conditions for 3-D reconstruction. The method focuses on the calibration of the extrinsic parameters which are subject to change since the effects of the workspace factors. The intrinsic parameters were calibrated in advance by an off-line method and were assumed as invariable. The results of simulation experiments demonstrate that the camera parameters could be calibrated effectively and achieve the real time need.

  4. Alternate calibration method of radiochromic EBT3 film for quality assurance verification of clinical radiotherapy treatments

    NASA Astrophysics Data System (ADS)

    Park, Soah; Kang, Sei-Kwon; Cheong, Kwang-Ho; Hwang, Taejin; Yoon, Jai-Woong; Koo, Taeryool; Han, Tae Jin; Kim, Haeyoung; Lee, Me Yeon; Bae, Hoonsik; Kim, Kyoung Ju

    2016-07-01

    EBT3 film is utilized as a dosimetry quality assurance tool for the verification of clinical radiotherapy treatments. In this work, we suggest a percentage-depth-dose (PDD) calibration method that can calibrate several EBT3 film pieces together at different dose levels because photon beams provide different dose levels at different depths along the axis of the beam. We investigated the feasibility of the film PDD calibration method based on PDD data and compared the results those from the traditional film calibration method. Photon beams at 6 MV were delivered to EBT3 film pieces for both calibration methods. For the PDD-based calibration, the film pieces were placed on solid phantoms at the depth of maximum dose (dmax) and at depths of 3, 5, 8, 12, 17, and 22 cm, and a photon beam was delivered twice, at 100 cGy and 400 cGy, to extend the calibration dose range under the same conditions. Fourteen film pieces, to maintain their consistency, were irradiated at doses ranging from approximately 30 to 400 cGy for both film calibrations. The film pieces were located at the center position on the scan bed of an Epson 1680 flatbed scanner in the parallel direction. Intensity-modulated radiation therapy (IMRT) plans were created, and their dose distributions were delivered to the film. The dose distributions for the traditional method and those for the PDD-based calibration method were evaluated using a Gamma analysis. The PDD dose values using a CC13 ion chamber and those obtained by using a FC65-G Farmer chamber and measured at the depth of interest produced very similar results. With the objective test criterion of a 1% dosage agreement at 1 mm, the passing rates for the four cases of the three IMRT plans were essentially identical. The traditional and the PDD-based calibrations provided similar plan verification results. We also describe another alternative for calibrating EBT3 films, i.e., a PDD-based calibration method that provides an easy and time-saving approach

  5. A new method for automated dynamic calibration of tipping-bucket rain gauges

    USGS Publications Warehouse

    Humphrey, M.D.; Istok, J.D.; Lee, J.Y.; Hevesi, J.A.; Flint, A.L.

    1997-01-01

    Existing methods for dynamic calibration of tipping-bucket rain gauges (TBRs) can be time consuming and labor intensive. A new automated dynamic calibration system has been developed to calibrate TBRs with minimal effort. The system consists of a programmable pump, datalogger, digital balance, and computer. Calibration is performed in two steps: 1) pump calibration and 2) rain gauge calibration. Pump calibration ensures precise control of water flow rates delivered to the rain gauge funnel; rain gauge calibration ensures precise conversion of bucket tip times to actual rainfall rates. Calibration of the pump and one rain gauge for 10 selected pump rates typically requires about 8 h. Data files generated during rain gauge calibration are used to compute rainfall intensities and amounts from a record of bucket tip times collected in the field. The system was tested using 5 types of commercial TBRs (15.2-, 20.3-, and 30.5-cm diameters; 0.1-, 0.2-, and 1.0-mm resolutions) and using 14 TBRs of a single type (20.3-cm diameter; 0.1-mm resolution). Ten pump rates ranging from 3 to 154 mL min-1 were used to calibrate the TBRs and represented rainfall rates between 6 and 254 mm h-1 depending on the rain gauge diameter. All pump calibration results were very linear with R2 values greater than 0.99. All rain gauges exhibited large nonlinear underestimation errors (between 5% and 29%) that decreased with increasing rain gauge resolution and increased with increasing rainfall rate, especially for rates greater than 50 mm h-1. Calibration curves of bucket tip time against the reciprocal of the true pump rate for all rain gauges also were linear with R2 values of 0.99. Calibration data for the 14 rain gauges of the same type were very similar, as indicated by slope values that were within 14% of each other and ranged from about 367 to 417 s mm h-1. The developed system can calibrate TBRs efficiently, accurately, and virtually unattended and could be modified for use with other

  6. a Method for Self-Calibration in Satellite with High Precision of Space Linear Array Camera

    NASA Astrophysics Data System (ADS)

    Liu, Wei; Qian, Fangming; Miao, Yuzhe; Wang, Rongjian

    2016-06-01

    At present, the on-orbit calibration of the geometric parameters of a space surveying camera is usually processed by data from a ground calibration field after capturing the images. The entire process is very complicated and lengthy and cannot monitor and calibrate the geometric parameters in real time. On the basis of a large number of on-orbit calibrations, we found that owing to the influence of many factors, e.g., weather, it is often difficult to capture images of the ground calibration field. Thus, regular calibration using field data cannot be ensured. This article proposes a real time self-calibration method for a space linear array camera on a satellite using the optical auto collimation principle. A collimating light source and small matrix array CCD devices are installed inside the load system of the satellite; these use the same light path as the linear array camera. We can extract the location changes of the cross marks in the matrix array CCD to determine the real-time variations in the focal length and angle parameters of the linear array camera. The on-orbit status of the camera is rapidly obtained using this method. On one hand, the camera's change regulation can be mastered accurately and the camera's attitude can be adjusted in a timely manner to ensure optimal photography; in contrast, self-calibration of the camera aboard the satellite can be realized quickly, which improves the efficiency and reliability of photogrammetric processing.

  7. A new method for internal calibration of nuclear track detectors

    NASA Technical Reports Server (NTRS)

    Oda, K.; Csige, I.; Henke, R. P.; Benton, E. V.

    1992-01-01

    A new technique is proposed for an internal calibration of a two-layer detector assembly. Spatially coincident pairs of conical tracks on one surface and overetched tracks on the adjacent surface are selected for measurement. Both the etch rate ratio and the particle range can be obtained from the minor and major diameters of the elliptical track and the radii of the circular tracks for two etching steps. This technique was applied to CR-39 detectors exposed to fast neutrons and those flown on a high altitude balloon in order to evaluate the proton response. An improvement by using multi-step etching was also carried out. It was found that not only a single set of the etch rate ratio and the range but also the response curve could be estimated in an extended region by analyzing combined growth curves.

  8. Calibration methods and tools for KM3NeT

    NASA Astrophysics Data System (ADS)

    Kulikovskiy, Vladimir

    2016-04-01

    The KM3NeT detectors, ARCA and ORCA, composed of several thousands digital optical modules, are in the process of their realization in the Mediterranean Sea. Each optical module contains 31 3-inch photomultipliers. Readout of the optical modules and other detector components is synchronized at the level of sub-nanoseconds. The position of the module is measured by acoustic piezo detectors inside the module and external acoustic emitters installed on the bottom of the sea. The orientation of the module is obtained with an internal attitude and heading reference system chip. Detector calibration, i.e. timing, positioning and sea-water properties, is overviewed in this talk and discussed in detail in this conference. Results of the procedure applied to the first detector unit ready for installation in the deep sea will be shown.

  9. A direct micropipette-based calibration method for atomic force microscope cantilevers

    PubMed Central

    Liu, Baoyu; Yu, Yan; Yao, Da-Kang; Shao, Jin-Yu

    2009-01-01

    In this report, we describe a direct method for calibrating atomic force microscope (AFM) cantilevers with the micropipette aspiration technique (MAT). A closely fitting polystyrene bead inside a micropipette is driven by precisely controlled hydrostatic pressures to apply known loads on the sharp tip of AFM cantilevers, thus providing a calibration at the most functionally relevant position. The new method is capable of calibrating cantilevers with spring constants ranging from 0.01 to hundreds of newtons per meter. Under appropriate loading conditions, this new method yields measurement accuracy and precision both within 10%, with higher performance for softer cantilevers. Furthermore, this method may greatly enhance the accuracy and precision of calibration for colloidal probes. PMID:19566228

  10. Nanoscale Calibration Standards and Methods: Dimensional and Related Measurements in the Micro- and Nanometer Range

    NASA Astrophysics Data System (ADS)

    Wilkening, Günter; Koenders, Ludger

    2005-08-01

    The quantitative determination of the properties of micro- and nanostructures is essential in research and development. It is also a prerequisite in process control and quality assurance in industry. The knowledge of the geometrical dimensions of structures in most cases is the base, to which other physical and chemical properties are linked. Quantitative measurements require reliable and stable instruments, suitable measurement procedures as well as appropriate calibration artefacts and methods. The seminar "NanoScale 2004" (6th Seminar on Quantitative Microscopy and 2nd Seminar on Nanoscale Calibration Standards and Methods) at the National Metrology Institute (Physikalisch-Technische Bundesanstalt PTB), Braunschweig, Germany, continues the series of seminars on Quantitative Microscopy. The series stimulates the exchange of information between manufacturers of relevant hard- and software and the users in science and industry. Topics addressed in these proceedings are a) the application of quantitative measurements and measurement problems in: microelectronics, microsystems technology, nano/quantum/molecular electronics, chemistry, biology, medicine, environmental technology, materials science, surface processing b) calibration & correction methods: calibration methods, calibration standards, calibration procedures, traceable measurements, standardization, uncertainty of measurements c) instrumentation and methods: novel/improved instruments and methods, reproducible probe/sample positioning, position-measuring systems, novel/improved probe/detector systems, linearization methods, image processing

  11. SERI results from the PEP 1987 Summit Round Robin and a comparison of photovoltaic calibration methods

    SciTech Connect

    Emery, K.A.; Waddington, D.; Rummel, S.; Myers, D.R.; Stoffel, T.L.; Osterwald, C.R.

    1989-03-01

    The SERI results for the Photovoltaic Solar Energy Project (PEP) 1987 summit round robin are presented and the proposed International Electrotechnical Commission (IEC) technical committee TC82 global calibration method is evaluated in this report. The global calibration method has a +-4% uncertainty in the short-circuit current (I/sub sc/) including random and nonrandom error sources. This uncertainty can be reduced +-3% if the global normal method is used and the direct component of the total irradiance is measured with a primary absolute cavity radiometer. Nearly half of the uncertainty for various primary and secondary PV calibration methods is nonrandom. If a +-4% uncertainty in I/sub sc/ under standard reporting conditions is acceptable, then the global calibration method with spectral corrections is suitable for cells or modules. If a primary reference device with an uncertainty less than +-3% is required, then the global method is unsuitable. The SERI primary direct normal calibration method has a +-1% uncertainty in I/sub sc/ as verified by comparison with primary AMO cells. If a +-1% primary reference cell is used for secondary calibrations, then a +-2% uncertainty could be achieved for reference cells or modules. 37 refs., 29 figs., 42 tabs.

  12. Quantitative Analysis of Carbon Steel with Multi-Line Internal Standard Calibration Method Using Laser-Induced Breakdown Spectroscopy.

    PubMed

    Pan, Congyuan; Du, Xuewei; An, Ning; Zeng, Qiang; Wang, Shengbo; Wang, Qiuping

    2016-04-01

    A multi-line internal standard calibration method is proposed for the quantitative analysis of carbon steel using laser-induced breakdown spectroscopy (LIBS). A procedure based on the method was adopted to select the best calibration curves and the corresponding emission lines pairs automatically. Laser-induced breakdown spectroscopy experiments with carbon steel samples were performed, and C, Cr, and Mn were analyzed via the proposed method. Calibration curves of these elements were constructed via a traditional single line internal standard calibration method and a multi-line internal standard calibration method. The calibration curves obtained were evaluated with the determination coefficient, the root mean square error of cross-validation, and the average relative error of cross-validation. All of the parameters were improved significantly with the proposed method. The results show that accurate and stable calibration curves can be obtained efficiently via the multi-line internal standard calibration method. PMID:26872822

  13. Novel Method for Processing the Dynamic Calibration Signal of Pressure Sensor

    PubMed Central

    Wang, Zhongyu; Li, Qiang; Wang, Zhuoran; Yan, Hu

    2015-01-01

    Dynamic calibration is one of the important ways to acquire the dynamic performance parameters of a pressure sensor. This research focuses on the processing method for the output of calibrated pressure sensor, and mainly attempts to solve the problem of extracting the true information of step response under strong interference noise. A dynamic calibration system based on a shock tube is established to excite the time-domain response signal of a calibrated pressure sensor. A key processing on difference modeling is applied for the obtained signal, and several generating sequences are established. A fusion process for the generating sequences is then undertaken, and the true information of the step response of the calibrated pressure sensor can be obtained. Finally, by implementing the common QR decomposition method to deal with the true information, a dynamic model characterizing the dynamic performance of the calibrated pressure sensor is established. A typical pressure sensor was used to perform calibration tests and a frequency-domain experiment for the sensor was also conducted. Results show that the proposed method could effectively filter strong interference noise in the output of the sensor and the corresponding dynamic model could effectively characterize the dynamic performance of the pressure sensor. PMID:26197324

  14. Automatic Calibration Method for Driver’s Head Orientation in Natural Driving Environment

    PubMed Central

    Fu, Xianping; Guan, Xiao; Peli, Eli; Liu, Hongbo; Luo, Gang

    2013-01-01

    Gaze tracking is crucial for studying driver’s attention, detecting fatigue, and improving driver assistance systems, but it is difficult in natural driving environments due to nonuniform and highly variable illumination and large head movements. Traditional calibrations that require subjects to follow calibrators are very cumbersome to be implemented in daily driving situations. A new automatic calibration method, based on a single camera for determining the head orientation and which utilizes the side mirrors, the rear-view mirror, the instrument board, and different zones in the windshield as calibration points, is presented in this paper. Supported by a self-learning algorithm, the system tracks the head and categorizes the head pose in 12 gaze zones based on facial features. The particle filter is used to estimate the head pose to obtain an accurate gaze zone by updating the calibration parameters. Experimental results show that, after several hours of driving, the automatic calibration method without driver’s corporation can achieve the same accuracy as a manual calibration method. The mean error of estimated eye gazes was less than 5°in day and night driving. PMID:24639620

  15. Novel Method for Processing the Dynamic Calibration Signal of Pressure Sensor.

    PubMed

    Wang, Zhongyu; Li, Qiang; Wang, Zhuoran; Yan, Hu

    2015-01-01

    Dynamic calibration is one of the important ways to acquire the dynamic performance parameters of a pressure sensor. This research focuses on the processing method for the output of calibrated pressure sensor, and mainly attempts to solve the problem of extracting the true information of step response under strong interference noise. A dynamic calibration system based on a shock tube is established to excite the time-domain response signal of a calibrated pressure sensor. A key processing on difference modeling is applied for the obtained signal, and several generating sequences are established. A fusion process for the generating sequences is then undertaken, and the true information of the step response of the calibrated pressure sensor can be obtained. Finally, by implementing the common QR decomposition method to deal with the true information, a dynamic model characterizing the dynamic performance of the calibrated pressure sensor is established. A typical pressure sensor was used to perform calibration tests and a frequency-domain experiment for the sensor was also conducted. Results show that the proposed method could effectively filter strong interference noise in the output of the sensor and the corresponding dynamic model could effectively characterize the dynamic performance of the pressure sensor. PMID:26197324

  16. Additives

    NASA Technical Reports Server (NTRS)

    Smalheer, C. V.

    1973-01-01

    The chemistry of lubricant additives is discussed to show what the additives are chemically and what functions they perform in the lubrication of various kinds of equipment. Current theories regarding the mode of action of lubricant additives are presented. The additive groups discussed include the following: (1) detergents and dispersants, (2) corrosion inhibitors, (3) antioxidants, (4) viscosity index improvers, (5) pour point depressants, and (6) antifouling agents.

  17. Transient Method for Determining Indoor Chemical Concentrations Based on SPME: Model Development and Calibration.

    PubMed

    Cao, Jianping; Xiong, Jianyin; Wang, Lixin; Xu, Ying; Zhang, Yinping

    2016-09-01

    Solid-phase microextraction (SPME) is regarded as a nonexhaustive sampling technique with a smaller extraction volume and a shorter extraction time than traditional sampling techniques and is hence widely used. The SPME sampling process is affected by the convection or diffusion effect along the coating surface, but this factor has seldom been studied. This paper derives an analytical model to characterize SPME sampling for semivolatile organic compounds (SVOCs) as well as for volatile organic compounds (VOCs) by considering the surface mass transfer process. Using this model, the chemical concentrations in a sample matrix can be conveniently calculated. In addition, the model can be used to determine the characteristic parameters (partition coefficient and diffusion coefficient) for typical SPME chemical samplings (SPME calibration). Experiments using SPME samplings of two typical SVOCs, dibutyl phthalate (DBP) in sealed chamber and di(2-ethylhexyl) phthalate (DEHP) in ventilated chamber, were performed to measure the two characteristic parameters. The experimental results demonstrated the effectiveness of the model and calibration method. Experimental data from the literature (VOCs sampled by SPME) were used to further validate the model. This study should prove useful for relatively rapid quantification of concentrations of different chemicals in various circumstances with SPME. PMID:27476381

  18. A method for in situ absolute DD yield calibration of neutron time-of-flight detectors on OMEGA using CR-39-based proton detectors

    SciTech Connect

    Waugh, C. J.; Rosenberg, M. J.; Zylstra, A. B.; Frenje, J. A.; Seguin, F. H.; Petrasso, R. D.; Glebov, V. Yu.; Sangster, T. C.; Stoeckl, C.

    2015-05-27

    Neutron time of flight (nTOF) detectors are used routinely to measure the absolute DD neutron yield at OMEGA. To check the DD yield calibration of these detectors, originally calibrated using indium activation systems, which in turn were cross-calibrated to NOVA nTOF detectors in the early 1990s, a direct in situ calibration method using CR-39 range filter proton detectors has been successfully developed. By measuring DD neutron and proton yields from a series of exploding pusher implosions at OMEGA, a yield calibration coefficient of 1.09 ± 0.02 (relative to the previous coefficient) was determined for the 3m nTOF detector. In addition, comparison of these and other shots indicates that significant reduction in charged particle flux anisotropies is achieved when bang time occurs significantly (on the order of 500 ps) after the trailing edge of the laser pulse. This is an important observation as the main source of the yield calibration error is due to particle anisotropies caused by field effects. The results indicate that the CR-39-nTOF in situ calibration method can serve as a valuable technique for calibrating and reducing the uncertainty in the DD absolute yield calibration of nTOF detector systems on OMEGA, the National Ignition Facility, and laser megajoule.

  19. A method for in situ absolute DD yield calibration of neutron time-of-flight detectors on OMEGA using CR-39-based proton detectors

    DOE PAGESBeta

    Waugh, C. J.; Rosenberg, M. J.; Zylstra, A. B.; Frenje, J. A.; Seguin, F. H.; Petrasso, R. D.; Glebov, V. Yu.; Sangster, T. C.; Stoeckl, C.

    2015-05-27

    Neutron time of flight (nTOF) detectors are used routinely to measure the absolute DD neutron yield at OMEGA. To check the DD yield calibration of these detectors, originally calibrated using indium activation systems, which in turn were cross-calibrated to NOVA nTOF detectors in the early 1990s, a direct in situ calibration method using CR-39 range filter proton detectors has been successfully developed. By measuring DD neutron and proton yields from a series of exploding pusher implosions at OMEGA, a yield calibration coefficient of 1.09 ± 0.02 (relative to the previous coefficient) was determined for the 3m nTOF detector. In addition,more » comparison of these and other shots indicates that significant reduction in charged particle flux anisotropies is achieved when bang time occurs significantly (on the order of 500 ps) after the trailing edge of the laser pulse. This is an important observation as the main source of the yield calibration error is due to particle anisotropies caused by field effects. The results indicate that the CR-39-nTOF in situ calibration method can serve as a valuable technique for calibrating and reducing the uncertainty in the DD absolute yield calibration of nTOF detector systems on OMEGA, the National Ignition Facility, and laser megajoule.« less

  20. A method for in situ absolute DD yield calibration of neutron time-of-flight detectors on OMEGA using CR-39-based proton detectors

    SciTech Connect

    Waugh, C. J. Zylstra, A. B.; Frenje, J. A.; Séguin, F. H.; Petrasso, R. D.; Rosenberg, M. J.; Glebov, V. Yu.; Sangster, T. C.; Stoeckl, C.

    2015-05-15

    Neutron time of flight (nTOF) detectors are used routinely to measure the absolute DD neutron yield at OMEGA. To check the DD yield calibration of these detectors, originally calibrated using indium activation systems, which in turn were cross-calibrated to NOVA nTOF detectors in the early 1990s, a direct in situ calibration method using CR-39 range filter proton detectors has been successfully developed. By measuring DD neutron and proton yields from a series of exploding pusher implosions at OMEGA, a yield calibration coefficient of 1.09 ± 0.02 (relative to the previous coefficient) was determined for the 3m nTOF detector. In addition, comparison of these and other shots indicates that significant reduction in charged particle flux anisotropies is achieved when bang time occurs significantly (on the order of 500 ps) after the trailing edge of the laser pulse. This is an important observation as the main source of the yield calibration error is due to particle anisotropies caused by field effects. The results indicate that the CR-39-nTOF in situ calibration method can serve as a valuable technique for calibrating and reducing the uncertainty in the DD absolute yield calibration of nTOF detector systems on OMEGA, the National Ignition Facility, and laser megajoule.

  1. A method for in situ absolute DD yield calibration of neutron time-of-flight detectors on OMEGA using CR-39-based proton detectors.

    PubMed

    Waugh, C J; Rosenberg, M J; Zylstra, A B; Frenje, J A; Séguin, F H; Petrasso, R D; Glebov, V Yu; Sangster, T C; Stoeckl, C

    2015-05-01

    Neutron time of flight (nTOF) detectors are used routinely to measure the absolute DD neutron yield at OMEGA. To check the DD yield calibration of these detectors, originally calibrated using indium activation systems, which in turn were cross-calibrated to NOVA nTOF detectors in the early 1990s, a direct in situ calibration method using CR-39 range filter proton detectors has been successfully developed. By measuring DD neutron and proton yields from a series of exploding pusher implosions at OMEGA, a yield calibration coefficient of 1.09 ± 0.02 (relative to the previous coefficient) was determined for the 3m nTOF detector. In addition, comparison of these and other shots indicates that significant reduction in charged particle flux anisotropies is achieved when bang time occurs significantly (on the order of 500 ps) after the trailing edge of the laser pulse. This is an important observation as the main source of the yield calibration error is due to particle anisotropies caused by field effects. The results indicate that the CR-39-nTOF in situ calibration method can serve as a valuable technique for calibrating and reducing the uncertainty in the DD absolute yield calibration of nTOF detector systems on OMEGA, the National Ignition Facility, and laser megajoule. PMID:26026524

  2. A method for in situ absolute DD yield calibration of neutron time-of-flight detectors on OMEGA using CR-39-based proton detectors

    NASA Astrophysics Data System (ADS)

    Waugh, C. J.; Rosenberg, M. J.; Zylstra, A. B.; Frenje, J. A.; Séguin, F. H.; Petrasso, R. D.; Glebov, V. Yu.; Sangster, T. C.; Stoeckl, C.

    2015-05-01

    Neutron time of flight (nTOF) detectors are used routinely to measure the absolute DD neutron yield at OMEGA. To check the DD yield calibration of these detectors, originally calibrated using indium activation systems, which in turn were cross-calibrated to NOVA nTOF detectors in the early 1990s, a direct in situ calibration method using CR-39 range filter proton detectors has been successfully developed. By measuring DD neutron and proton yields from a series of exploding pusher implosions at OMEGA, a yield calibration coefficient of 1.09 ± 0.02 (relative to the previous coefficient) was determined for the 3m nTOF detector. In addition, comparison of these and other shots indicates that significant reduction in charged particle flux anisotropies is achieved when bang time occurs significantly (on the order of 500 ps) after the trailing edge of the laser pulse. This is an important observation as the main source of the yield calibration error is due to particle anisotropies caused by field effects. The results indicate that the CR-39-nTOF in situ calibration method can serve as a valuable technique for calibrating and reducing the uncertainty in the DD absolute yield calibration of nTOF detector systems on OMEGA, the National Ignition Facility, and laser megajoule.

  3. A Tool-Free Calibration Method for Turntable-Based 3D Scanning Systems.

    PubMed

    Pang, Xufang; Lau, Rynson W H; Song, Zhan; Li, Yangyan; He, Shengfeng

    2016-01-01

    Turntable-based 3D scanners are popular but require calibration of the turntable axis. Existing methods for turntable calibration typically make use of specially designed tools, such as a chessboard or criterion sphere, which users must manually install and dismount. In this article, the authors propose an automatic method to calibrate the turntable axis without any calibration tools. Given a scan sequence of the input object, they first recover the initial rotation axis from an automatic registration step. Then they apply an iterative procedure to obtain the optimized turntable axis. This iterative procedure alternates between two steps: refining the initial pose of the input scans and approximating the rotation matrix. The performance of the proposed method was evaluated on a structured light-based scanning system. PMID:25137724

  4. Automated calibration methods for robotic multisensor landmine detection

    NASA Astrophysics Data System (ADS)

    Keranen, Joe G.; Miller, Jonathan; Schultz, Gregory; Topolosky, Zeke

    2007-04-01

    Both force protection and humanitarian demining missions require efficient and reliable detection and discrimination of buried anti-tank and anti-personnel landmines. Widely varying surface and subsurface conditions, mine types and placement, as well as environmental regimes challenge the robustness of the automatic target recognition process. In this paper we present applications created for the U.S. Army Nemesis detection platform. Nemesis is an unmanned rubber-tracked vehicle-based system designed to eradicate a wide variety of anti-tank and anti-personnel landmines for humanitarian demining missions. The detection system integrates advanced ground penetrating synthetic aperture radar (GPSAR) and electromagnetic induction (EMI) arrays, highly accurate global and local positioning, and on-board target detection/classification software on the front loader of a semi-autonomous UGV. An automated procedure is developed to estimate the soil's dielectric constant using surface reflections from the ground penetrating radar. The results have implications not only for calibration of system data acquisition parameters, but also for user awareness and tuning of automatic target recognition detection and discrimination algorithms.

  5. GMI Instrument Spin Balance Method, Optimization, Calibration and Test

    NASA Technical Reports Server (NTRS)

    Ayari, Laoucet; Kubitschek, Michael; Ashton, Gunnar; Johnston, Steve; Debevec, Dave; Newell, David; Pellicciotti, Joseph

    2014-01-01

    The Global Microwave Imager (GMI) instrument must spin at a constant rate of 32 rpm continuously for the 3-year mission life. Therefore, GMI must be very precisely balanced about the spin axis and center of gravity (CG) to maintain stable scan pointing and to minimize disturbances imparted to the spacecraft and attitude control on-orbit. The GMI instrument is part of the core Global Precipitation Measurement (GPM) spacecraft and is used to make calibrated radiometric measurements at multiple microwave frequencies and polarizations. The GPM mission is an international effort managed by the National Aeronautics and Space Administration (NASA) to improve climate, weather, and hydro-meteorological predictions through more accurate and frequent precipitation measurements. Ball Aerospace and Technologies Corporation (BATC) was selected by NASA Goddard Space Flight Center to design, build, and test the GMI instrument. The GMI design has to meet a challenging set of spin balance requirements and had to be brought into simultaneous static and dynamic spin balance after the entire instrument was already assembled and before environmental tests began. The focus of this contribution is on the analytical and test activities undertaken to meet the challenging spin balance requirements of the GMI instrument. The novel process of measuring the residual static and dynamic imbalances with a very high level of accuracy and precision is presented together with the prediction of the optimal balance masses and their locations.

  6. GMI Instrument Spin Balance Method, Optimization, Calibration, and Test

    NASA Technical Reports Server (NTRS)

    Ayari, Laoucet; Kubitschek, Michael; Ashton, Gunnar; Johnston, Steve; Debevec, Dave; Newell, David; Pellicciotti, Joseph

    2014-01-01

    The Global Microwave Imager (GMI) instrument must spin at a constant rate of 32 rpm continuously for the 3 year mission life. Therefore, GMI must be very precisely balanced about the spin axis and CG to maintain stable scan pointing and to minimize disturbances imparted to the spacecraft and attitude control on-orbit. The GMI instrument is part of the core Global Precipitation Measurement (GPM) spacecraft and is used to make calibrated radiometric measurements at multiple microwave frequencies and polarizations. The GPM mission is an international effort managed by the National Aeronautics and Space Administration (NASA) to improve climate, weather, and hydro-meteorological predictions through more accurate and frequent precipitation measurements. Ball Aerospace and Technologies Corporation (BATC) was selected by NASA Goddard Space Flight Center to design, build, and test the GMI instrument. The GMI design has to meet a challenging set of spin balance requirements and had to be brought into simultaneous static and dynamic spin balance after the entire instrument was already assembled and before environmental tests began. The focus of this contribution is on the analytical and test activities undertaken to meet the challenging spin balance requirements of the GMI instrument. The novel process of measuring the residual static and dynamic imbalances with a very high level of accuracy and precision is presented together with the prediction of the optimal balance masses and their locations.

  7. Single Vector Calibration System for Multi-Axis Load Cells and Method for Calibrating a Multi-Axis Load Cell

    NASA Technical Reports Server (NTRS)

    Parker, Peter A. (Inventor)

    2003-01-01

    A single vector calibration system is provided which facilitates the calibration of multi-axis load cells, including wind tunnel force balances. The single vector system provides the capability to calibrate a multi-axis load cell using a single directional load, for example loading solely in the gravitational direction. The system manipulates the load cell in three-dimensional space, while keeping the uni-directional calibration load aligned. The use of a single vector calibration load reduces the set-up time for the multi-axis load combinations needed to generate a complete calibration mathematical model. The system also reduces load application inaccuracies caused by the conventional requirement to generate multiple force vectors. The simplicity of the system reduces calibration time and cost, while simultaneously increasing calibration accuracy.

  8. Radiometric calibration method for large aperture infrared system with broad dynamic range.

    PubMed

    Sun, Zhiyuan; Chang, Songtao; Zhu, Wei

    2015-05-20

    Infrared radiometric measurements can acquire important data for missile defense systems. When observation is carried out by ground-based infrared systems, a missile is characterized by long distance, small size, and large variation of radiance. Therefore, the infrared systems should be manufactured with a larger aperture to enhance detection ability and calibrated at a broader dynamic range to extend measurable radiance. Nevertheless, the frequently used calibration methods demand an extended-area blackbody with broad dynamic range or a huge collimator for filling the system's field stop, which would greatly increase manufacturing costs and difficulties. To overcome this restriction, a calibration method based on amendment of inner and outer calibration is proposed. First, the principles and procedures of this method are introduced. Then, a shifting strategy of infrared systems for measuring targets with large fluctuations of infrared radiance is put forward. Finally, several experiments are performed on a shortwave infrared system with Φ400  mm aperture. The results indicate that the proposed method cannot only ensure accuracy of calibration but have the advantage of low cost, low power, and high motility. Hence, it is an effective radiometric calibration method in the outfield. PMID:26192499

  9. A flexile and high precision calibration method for binocular structured light scanning system.

    PubMed

    Yuan, Jianying; Wang, Qiong; Li, Bailin

    2014-01-01

    3D (three-dimensional) structured light scanning system is widely used in the field of reverse engineering, quality inspection, and so forth. Camera calibration is the key for scanning precision. Currently, 2D (two-dimensional) or 3D fine processed calibration reference object is usually applied for high calibration precision, which is difficult to operate and the cost is high. In this paper, a novel calibration method is proposed with a scale bar and some artificial coded targets placed randomly in the measuring volume. The principle of the proposed method is based on hierarchical self-calibration and bundle adjustment. We get initial intrinsic parameters from images. Initial extrinsic parameters in projective space are estimated with the method of factorization and then upgraded to Euclidean space with orthogonality of rotation matrix and rank 3 of the absolute quadric as constraint. Last, all camera parameters are refined through bundle adjustment. Real experiments show that the proposed method is robust, and has the same precision level as the result using delicate artificial reference object, but the hardware cost is very low compared with the current calibration method used in 3D structured light scanning system. PMID:25202736

  10. Additive manufacturing method for SRF components of various geometries

    DOEpatents

    Rimmer, Robert; Frigola, Pedro E; Murokh, Alex Y

    2015-05-05

    An additive manufacturing method for forming nearly monolithic SRF niobium cavities and end group components of arbitrary shape with features such as optimized wall thickness and integral stiffeners, greatly reducing the cost and technical variability of conventional cavity construction. The additive manufacturing method for forming an SRF cavity, includes atomizing niobium to form a niobium powder, feeding the niobium powder into an electron beam melter under a vacuum, melting the niobium powder under a vacuum in the electron beam melter to form an SRF cavity; and polishing the inside surface of the SRF cavity.

  11. Vicarious calibration of the Ocean PHILLS hyperspectral sensor using a coastal tree-shadow method

    NASA Astrophysics Data System (ADS)

    Filippi, Anthony M.; Carder, Kendall L.; Davis, Curtiss O.

    2006-11-01

    Ocean color remote-sensing systems require highly accurate calibration (<0.5%) for accurate retrieval of water properties. This accuracy is typically achieved by vicarious calibration which is done by comparing the atmospherically corrected remote-sensing data to accurate estimates of the water-leaving radiance. Here we present a new method for vicarious calibration of a hyperspectral sensor that exploits shadows cast by trees and cliffs along coastlines. Hyperspectral Ocean PHILLS imagery was acquired over East Sound and adjacent waters around Orcas Island, Washington, USA, in August, 1998, in concert with field data collection. To vicariously calibrate the PHILLS data, a method was developed employing pixel pairs in tree-shaded and adjacent unshadowed waters, which utilizes the sky radiance dominating the shaded pixel as a known calibration target. Transects extracted from East Sound imagery were calibrated and validated with field data (RMSE = 0.00033 sr-1), providing validation of this approach for acquiring calibration-adjustment data from the image itself.

  12. A new automated and precise calibration method for gamma level gauges with rod detector arrangement.

    PubMed

    Peyvandi, Reza Gholipour; Taheri, Ali; Olfateh, Ali; Islami, Seyyedeh Zahra

    2016-06-01

    Gamma-ray liquid level gauging is of particular importance in several industries. Industrial vessels, tanks, and reactors, which work at high temperatures and pressures, usually have thick metal walls up to 20cm. These factors make it impossible to know the exact level of liquid or fluid while the system is operating. For this reason, the calibration process of the gamma level gauges is difficult as it is impossible to gain access to the inside of the vessels, which is important during the calibration process. In this study, a new auto-calibration method was proposed for the aforementioned situations. PMID:26974485

  13. Balance Calibration – A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance.

    SciTech Connect

    Mike Stears

    2010-07-01

    Paper Title: Balance Calibration – A method for assigning a direct-reading uncertainty to an electronic balance. Intended Audience: Those who calibrate or use electronic balances. Abstract: As a calibration facility, we provide on-site (at the customer’s location) calibrations of electronic balances for customers within our company. In our experience, most of our customers are not using their balance as a comparator, but simply putting an unknown quantity on the balance and reading the displayed mass value. Manufacturer’s specifications for balances typically include specifications such as readability, repeatability, linearity, and sensitivity temperature drift, but what does this all mean when the balance user simply reads the displayed mass value and accepts the reading as the true value? This paper discusses a method for assigning a direct-reading uncertainty to a balance based upon the observed calibration data and the environment where the balance is being used. The method requires input from the customer regarding the environment where the balance is used and encourages discussion with the customer regarding sources of uncertainty and possible means for improvement; the calibration process becomes an educational opportunity for the balance user as well as calibration personnel. This paper will cover the uncertainty analysis applied to the calibration weights used for the field calibration of balances; the uncertainty is calculated over the range of environmental conditions typically encountered in the field and the resulting range of air density. The temperature stability in the area of the balance is discussed with the customer and the temperature range over which the balance calibration is valid is decided upon; the decision is based upon the uncertainty needs of the customer and the desired rigor in monitoring by the customer. Once the environmental limitations are decided, the calibration is performed and the measurement data is entered into a

  14. A flexible method for calibrating external parameters of two cameras with no-overlapping FOV

    NASA Astrophysics Data System (ADS)

    Shao, Mingwei; Wei, Zhenzhong; Hu, Mengjie

    2016-01-01

    A new flexible method to calibrate the external parameters of two cameras with no-overlapping field of view (FOV) is proposed in our paper. A flexible target with four spheres and a 1D bar is designed. All spheres can move freely along the bar to make sure that each camera can capture the image of two spheres clearly. As the radius of each sphere is known exactly, the center of each sphere under its corresponding camera coordinate system can be confirmed from each sphere projection. The centers of the four spheres are collinear in the process of calibration, so we can express the relationship of the four centers only by external parameters of the two cameras. When the expressions in different positions are obtained, the external parameters of two cameras can be determined. In our proposed calibration method, the center of the sphere can be determined accurately as the sphere projection is not concerned with the sphere orientation, meanwhile, the freely movement of the spheres can ensure the image of spheres clearly. Experiment results show that the proposed calibration method can obtain an acceptable accuracy, the calibrated vision system reaches 0.105 mm when measuring a distance section of 1040 mm. Moreover, the calibration method is efficient, convenient and with an easy operation.

  15. A novel separation and calibration method for DVL and compass error in dead reckoning navigation systems

    NASA Astrophysics Data System (ADS)

    Zhang, Yanshun; Guo, Yajing; Yang, Tao; Li, Chunyu; Wang, Zhanqing

    2016-06-01

    The scale factor error δ C of the Doppler velocity log (DVL) and the heading angle error δ \\psi of a compass are so integrated in dead reckoning (DR) navigation systems that it is difficult to separate them. This paper aims to solve this problem by putting forward an online separation and calibration method for δ C and δ \\psi based on an ‘arc and linear’ trajectory. This method introduces the high-accuracy location information of a long base line (LBL) acoustic positioning system. At first, the relationship between the displacements on the ‘arc’ trajectory in directions of east and north, output by the LBL and DR systems, serves to judge the carrier direction and calibrate δ C . And then by compensating δ C , the displacement on the ‘linear’ trajectory is used to calibrate δ \\psi . Finally, a semi-physical simulation experiment is conducted to test and verify this calibration method to see how effective and accurate it is. Experimental results show that after calibration the residual error ratios of δ C and δ \\psi are 8.24% and 3.70% respectively. Therefore, online calibration of δ C and δ \\psi is realized effectively. What’s more, when the DR system is working alone in 400 s, this method reduces position error by up to 93.39%, from 18.91 m to 1.25 m.

  16. A new method for the absolute radiance calibration for UV-vis measurements of scattered sunlight

    NASA Astrophysics Data System (ADS)

    Wagner, T.; Beirle, S.; Dörner, S.; Penning de Vries, M.; Remmers, J.; Rozanov, A.; Shaiganfar, R.

    2015-10-01

    Absolute radiometric calibrations are important for measurements of the atmospheric spectral radiance. Such measurements can be used to determine actinic fluxes, the properties of aerosols and clouds, and the shortwave energy budget. Conventional calibration methods in the laboratory are based on calibrated light sources and reflectors and are expensive, time consuming and subject to relatively large uncertainties. Also, the calibrated instruments might change during transport from the laboratory to the measurement sites. Here we present a new calibration method for UV-vis instruments that measure the spectrally resolved sky radiance, for example zenith sky differential optical absorption spectroscopy (DOAS) instruments or multi-axis (MAX)-DOAS instruments. Our method is based on the comparison of the solar zenith angle dependence of the measured zenith sky radiance with radiative transfer simulations. For the application of our method, clear-sky measurements during periods with almost constant aerosol optical depth are needed. The radiative transfer simulations have to take polarisation into account. We show that the calibration results are almost independent from the knowledge of the aerosol optical properties and surface albedo, which causes a rather small uncertainty of about < 7 %. For wavelengths below about 330 nm it is essential that the ozone column density during the measurements be constant and known.

  17. A Radial Self-Calibrated (RASCAL) GRAPPA method using Weight Interpolation

    PubMed Central

    Codella, Noel C. F.; Spincemaille, Pascal; Prince, Martin; Wang, Yi

    2011-01-01

    A generalized autocalibrating partially parallel acquisition (GRAPPA) method for radial k-space sampling is presented that calculates GRAPPA weights without synthesized or acquired calibration data. Instead, GRAPPA weights are fit to the undersampled data as if it were the calibration data itself. Because the relative k-space shifts associated with these GRAPPA weights are varying for a radial trajectory, new GRAPPA weights can be resampled for arbitrary shifts through interpolation, which is then used to generate missing projections between the acquired projections. The method is demonstrated in phantoms and in abdominal and brain imaging. Image quality is similar to radial GRAPPA using fully sampled calibration data, and improved compared to a previously described self-calibrated radial GRAPPA technique. PMID:21834008

  18. A new and simple calibration-independent method for measuring the beam energy of a cyclotron.

    PubMed

    Gagnon, Katherine; Jensen, Mikael; Thisgaard, Helge; Publicover, Julia; Lapi, Suzanne; McQuarrie, Steve A; Ruth, Thomas J

    2011-01-01

    This work recommends a new and simple-to-perform method for measuring the beam energy of an accelerator. The proposed method requires the irradiation of two monitor foils interspaced by an energy degrader. The primary advantage of the proposed method, which makes this method unique from previous energy evaluation strategies that employ the use of monitor foils, is that this method is independent of the detector efficiency calibration. This method was evaluated by performing proton activation of (nat)Cu foils using both a cyclotron and a tandem Van de Graaff accelerator. The monitor foil activities were read using a dose calibrator set to an arbitrary calibration setting. Excellent agreement was noted between the nominal and measured proton energies. PMID:20926304

  19. New methods of data calibration for high power-aperture lidar.

    PubMed

    Guan, Sai; Yang, Guotao; Chang, Qihai; Cheng, Xuewu; Yang, Yong; Gong, Shaohua; Wang, Jihong

    2013-03-25

    For high power-aperture lidar sounding of wide atmospheric dynamic ranges, as in middle-upper atmospheric probing, photomultiplier tubes' (PMT) pulse pile-up effects and signal-induced noise (SIN) complicates the extraction of information from lidar return signal, especially from metal layers' fluorescence signal. Pursuit for sophisticated description of metal layers' characteristics at far range (80~130km) with one PMT of high quantum efficiency (QE) and good SNR, contradicts the requirements for signals of wide linear dynamic range (i.e. from approximate 10(2) to 10(8) counts/s). In this article, Substantial improvements on experimental simulation of Lidar signals affected by PMT are reported to evaluate the PMTs' distortions in our High Power-Aperture Sodium LIDAR system. A new method for pile-up calibration is proposed by taking into account PMT and High Speed Data Acquisition Card as an Integrated Black-Box, as well as a new experimental method for identifying and removing SIN from the raw Lidar signals. Contradiction between the limited linear dynamic range of raw signal (55~80km) and requirements for wider acceptable linearity has been effectively solved, without complicating the current lidar system. Validity of these methods was demonstrated by applying calibrated data to retrieve atmospheric parameters (i.e. atmospheric density, temperature and sodium absolutely number density), in comparison with measurements of TIMED satellite and atmosphere model. Good agreements are obtained between results derived from calibrated signal and reference measurements where differences of atmosphere density, temperature are less than 5% in the stratosphere and less than 10K from 30km to mesosphere, respectively. Additionally, approximate 30% changes are shown in sodium concentration at its peak value. By means of the proposed methods to revert the true signal independent of detectors, authors approach a new balance between maintaining the linearity of adequate signal (20

  20. Calibration of the analogue method for precipitation forecasting by means of genetic algorithms

    NASA Astrophysics Data System (ADS)

    Horton, P.; Jaboyedoff, M.; Obled, C.

    2012-04-01

    The analogue downscaling technique allows precipitation forecasting on the basis of the synoptic circulation and humidity variables resulting from a global circulation model (GCM). The method identifies analog days in a long archive of past situations and uses their observed precipitation amount to build the empirical conditional distribution considered as the probabilistic forecast for the target day. The Atmoswing model (Analog Technique MOdel for Statistical Weather forecastING) was developed to calibrate the method and to process real-time forecasting in the Swiss Alps. It is part of the MINERVE project, which aims at reducing the flood peaks of the Rhône River by means of water retention in dams. Such a method is highly non-linear, works with both discrete and continuous variables and has a complex cost surface. Calibration with linear methods such as a simplex concept has been tried, but has led to unsatisfying results. The complexity of the analogue technique gives the user no choice but to use either a step-by-step manual calibration or a global optimizer. While the first option is the commonly used approach, a global optimizer has never been used to fulfill that goal. The classic calibration's main issue is that parameters are not independent and the choices made in the beginning of the calibration procedure have an impact on the final set. We decided to implement the genetic algorithms to achieve the model calibration, and so to avoid subjective choices of initial parameters. Genetic algorithms exist with multiple operators (natural selection, mating selection, chromosomes crossover, and mutation) variations and specific parameterizations. Most used implementations were compared to choose the version resulting in the best calibration with the minimum processing time. Afterwards, the optimizer is used to explore new variable spaces and even to choose the best atmospheric variables, what was not possible in the traditional calibration procedure.

  1. Improvements to and comparison of static terrestrial LiDAR self-calibration methods.

    PubMed

    Chow, Jacky C K; Lichti, Derek D; Glennie, Craig; Hartzell, Preston

    2013-01-01

    Terrestrial laser scanners are sophisticated instruments that operate much like high-speed total stations. It has previously been shown that unmodelled systematic errors can exist in modern terrestrial laser scanners that deteriorate their geometric measurement precision and accuracy. Typically, signalised targets are used in point-based self-calibrations to identify and model the systematic errors. Although this method has proven its effectiveness, a large quantity of signalised targets is required and is therefore labour-intensive and limits its practicality. In recent years, feature-based self-calibration of aerial, mobile terrestrial, and static terrestrial laser scanning systems has been demonstrated. In this paper, the commonalities and differences between point-based and plane-based self-calibration (in terms of model identification and parameter correlation) are explored. The results of this research indicate that much of the knowledge from point-based self-calibration can be directly transferred to plane-based calibration and that the two calibration approaches are nearly equivalent. New network configurations, such as the inclusion of tilted scans, were also studied and prove to be an effective means for strengthening the self-calibration solution, and improved recoverability of the horizontal collimation axis error for hybrid scanners, which has always posed a challenge in the past. PMID:23727956

  2. Improvements to and Comparison of Static Terrestrial LiDAR Self-Calibration Methods

    PubMed Central

    Chow, Jacky C. K.; Lichti, Derek D.; Glennie, Craig; Hartzell, Preston

    2013-01-01

    Terrestrial laser scanners are sophisticated instruments that operate much like high-speed total stations. It has previously been shown that unmodelled systematic errors can exist in modern terrestrial laser scanners that deteriorate their geometric measurement precision and accuracy. Typically, signalised targets are used in point-based self-calibrations to identify and model the systematic errors. Although this method has proven its effectiveness, a large quantity of signalised targets is required and is therefore labour-intensive and limits its practicality. In recent years, feature-based self-calibration of aerial, mobile terrestrial, and static terrestrial laser scanning systems has been demonstrated. In this paper, the commonalities and differences between point-based and plane-based self-calibration (in terms of model identification and parameter correlation) are explored. The results of this research indicate that much of the knowledge from point-based self-calibration can be directly transferred to plane-based calibration and that the two calibration approaches are nearly equivalent. New network configurations, such as the inclusion of tilted scans, were also studied and prove to be an effective means for strengthening the self-calibration solution, and improved recoverability of the horizontal collimation axis error for hybrid scanners, which has always posed a challenge in the past. PMID:23727956

  3. Local Strategy Combined with a Wavelength Selection Method for Multivariate Calibration.

    PubMed

    Chang, Haitao; Zhu, Lianqing; Lou, Xiaoping; Meng, Xiaochen; Guo, Yangkuan; Wang, Zhongyu

    2016-01-01

    One of the essential factors influencing the prediction accuracy of multivariate calibration models is the quality of the calibration data. A local regression strategy, together with a wavelength selection approach, is proposed to build the multivariate calibration models based on partial least squares regression. The local algorithm is applied to create a calibration set of spectra similar to the spectrum of an unknown sample; the synthetic degree of grey relation coefficient is used to evaluate the similarity. A wavelength selection method based on simple-to-use interactive self-modeling mixture analysis minimizes the influence of noisy variables, and the most informative variables of the most similar samples are selected to build the multivariate calibration model based on partial least squares regression. To validate the performance of the proposed method, ultraviolet-visible absorbance spectra of mixed solutions of food coloring analytes in a concentration range of 20-200 µg/mL is measured. Experimental results show that the proposed method can not only enhance the prediction accuracy of the calibration model, but also greatly reduce its complexity. PMID:27271636

  4. Induced Dual-Nanospray: A Novel Internal Calibration Method for Convenient and Accurate Mass Measurement

    NASA Astrophysics Data System (ADS)

    Li, Yafeng; Zhang, Ning; Zhou, Yueming; Wang, Jianing; Zhang, Yiming; Wang, Jiyun; Xiong, Caiqiao; Chen, Suming; Nie, Zongxiu

    2013-09-01

    Accurate mass information is of great importance in the determination of unknown compounds. An effective and easy-to-control internal mass calibration method will dramatically benefit accurate mass measurement. Here we reported a simple induced dual-nanospray internal calibration device which has the following three advantages: (1) the two sprayers are in the same alternating current field; thus both reference ions and sample ions can be simultaneously generated and recorded. (2) It is very simple and can be easily assembled. Just two metal tubes, two nanosprayers, and an alternating current power supply are included. (3) With the low-flow-rate character and the versatility of nanoESI, this calibration method is capable of calibrating various samples, even untreated complex samples such as urine and other biological samples with small sample volumes. The calibration errors are around 1 ppm in positive ion mode and 3 ppm in negative ion mode with good repeatability. This new internal calibration method opens up new possibilities in the determination of unknown compounds, and it has great potential for the broad applications in biological and chemical analysis.

  5. A Visual Servoing-Based Method for ProCam Systems Calibration

    PubMed Central

    Berry, Francois; Aider, Omar Ait; Mosnier, Jeremie

    2013-01-01

    Projector-camera systems are currently used in a wide field of applications, such as 3D reconstruction and augmented reality, and can provide accurate measurements, depending on the configuration and calibration. Frequently, the calibration task is divided into two steps: camera calibration followed by projector calibration. The latter still poses certain problems that are not easy to solve, such as the difficulty in obtaining a set of 2D–3D points to compute the projection matrix between the projector and the world. Existing methods are either not sufficiently accurate or not flexible. We propose an easy and automatic method to calibrate such systems that consists in projecting a calibration pattern and superimposing it automatically on a known printed pattern. The projected pattern is provided by a virtual camera observing a virtual pattern in an OpenGL model. The projector displays what the virtual camera visualizes. Thus, the projected pattern can be controlled and superimposed on the printed one with the aid of visual servoing. Our experimental results compare favorably with those of other methods considering both usability and accuracy. PMID:24084121

  6. Local Strategy Combined with a Wavelength Selection Method for Multivariate Calibration

    PubMed Central

    Chang, Haitao; Zhu, Lianqing; Lou, Xiaoping; Meng, Xiaochen; Guo, Yangkuan; Wang, Zhongyu

    2016-01-01

    One of the essential factors influencing the prediction accuracy of multivariate calibration models is the quality of the calibration data. A local regression strategy, together with a wavelength selection approach, is proposed to build the multivariate calibration models based on partial least squares regression. The local algorithm is applied to create a calibration set of spectra similar to the spectrum of an unknown sample; the synthetic degree of grey relation coefficient is used to evaluate the similarity. A wavelength selection method based on simple-to-use interactive self-modeling mixture analysis minimizes the influence of noisy variables, and the most informative variables of the most similar samples are selected to build the multivariate calibration model based on partial least squares regression. To validate the performance of the proposed method, ultraviolet-visible absorbance spectra of mixed solutions of food coloring analytes in a concentration range of 20–200 µg/mL is measured. Experimental results show that the proposed method can not only enhance the prediction accuracy of the calibration model, but also greatly reduce its complexity. PMID:27271636

  7. Data Sparseness and On-Line Pretest Item Calibration-Scaling Methods in CAT.

    ERIC Educational Resources Information Center

    Ban, Jae-Chun; Hanson, Bradley A.; Yi, Qing; Harris, Deborah J.

    2002-01-01

    Compared three online pretest calibration scaling methods through simulation: (1) marginal maximum likelihood with one expectation maximization (EM) cycle (OEM) method; (2) marginal maximum likelihood with multiple EM cycles (MEM); and (3) M. Stocking's method B. MEM produced the smallest average total error in parameter estimation; OEM yielded…

  8. Data Sparseness and Online Pretest Item Calibration/Scaling Methods in CAT. ACT Research Report Series.

    ERIC Educational Resources Information Center

    Ban, Jae-Chun; Hanson, Bradley A.; Yi, Qing; Harris, Deborah J.

    The purpose of this study was to compare and evaluate three online pretest item calibration/scaling methods in terms of item parameter recovery when the item responses to the pretest items in the pool would be sparse. The three methods considered were the marginal maximum likelihood estimate with one EM cycle (OEM) method, the marginal maximum…

  9. Gafchromic EBT2 film dosimetry in reflection mode with a novel plan-based calibration method

    SciTech Connect

    Mendez, I.; Hartman, V.; Hudej, R.; Strojnik, A.; Casar, B.

    2013-01-15

    Purpose:A dosimetric system formed by Gafchromic EBT2 radiochromic film and Epson Expression 10000XL flatbed scanner was commissioned for dosimetry. In this paper, several open questions concerning the commissioning of radiochromic films for dosimetry were addressed: (a) is it possible to employ this dosimetric system in reflection mode; (b) if so, can the methods used in transmission mode also be used in reflection mode; (c) is it possible to obtain accurate absolute dose measurements with Gafchromic EBT2 films; (d) which calibration method should be followed; (e) which calibration models should be used; and (f) does three-color channel dosimetry offer a significant improvement over single channel dosimetry. The purpose of this paper is to help clarify these questions. Methods: In this study, films were scanned in reflection mode, the effect of surrounding film was evaluated and the feasibility of EBT2 film dosimetry in reflection mode was studied. EBT2's response homogeneity has been reported to lead to excessive dose uncertainties. To overcome this problem, a new plan-based calibration method was implemented. Plan-based calibration can use every pixel and each of the three color channels of the scanned film to obtain the parameters of the calibration model. A model selection analysis was conducted to select lateral correction and sensitometric curve models. The commonly used calibration with fragments was compared with red-channel plan-based calibration and with three-channel plan-based calibration. Results: No effect of surrounding film was found in this study. The film response inhomogeneity in EBT2 films was found to be important not only due to differences in the fog but also due to differences in sensitivity. The best results for lateral corrections were obtained using absolute corrections independent of the dose. With respect to the sensitometric curves, an empirical polynomial fit of order 4 was found to obtain results equivalent to a gamma

  10. Calibration of Hydraulic Conductivities by the Kalman Filtered Double Constraint Method

    NASA Astrophysics Data System (ADS)

    Zijl, Wouter; El-Rawy, Mustafa; Batelaan, Okke

    2014-05-01

    To assess the consequences of a changing environment for future management decisions we need quantitative techniques validated by case studies. In this context dealing with the limited data availability and inherent uncertainty is a major challenge. In this contribution we present a combination of two techniques (the Double Constraint Method and the Kalman Filter) exemplified by case studies. The techniques assist in the calibration of hydraulic grid block conductivities as well as in finding the reliability of the result. To focus on the basic principles we exemplify our approach for flow in which storage by water compressibility and pore space deformation is negligible. Only storage by water table movements plays a role. Such conditions hold for most flow problems in the relatively shallow aquifer-aquitard systems of deltaic regions. In a forward problem the conductivity is specified in all grid blocks. In addition, on each point of the boundary and in each well only one type of boundary condition has to be specified: either head, or flux. Our approach is based on the principle that calibration of the initial conductivities is meaningful only if we can specify both head and flux at a number of boundary points or wells, including no-flux monitoring wells. In general a hydrogeological model is a "flux model," i.e., the model is as much as possible based on specified ("measured") fluxes through the boundaries and in the wells. An exception is Tóth's flow systems analysis where, instead of the usual recharge fluxes, heads are specified on the water table. This suggests building a second forward model, a "head model," that is as much as possible based on specified (measured) heads on the boundaries and in the wells. The initial conductivities are then updated by applying Darcy's law K = -q/(¶h/¶x) to the fluxes q obtained by the "flux model" and the head gradients ¶h/¶x obtained by the "head model." This so-called "double constraint method" (DCM) leads to a

  11. Technical note: comparing calibration methods for determination of protein in goat milk by ultraviolet spectroscopy.

    PubMed

    Rukke, E O; Olsen, E F; Devold, T; Vegarud, G; Isaksson, T

    2010-07-01

    A rapid spectroscopic method to determine total protein in bovine and buffalo milk using UV spectra of guanidine-hydrochloride mixed milk has previously been reported and validated. The method was based on mixed calibration samples and univariate calibrations of fourth derivative (4D) spectra. In this study the same method was compared and tested for determination of total protein in goat milk. Calculations based on multivariate calibration (partial least squares regression) on full spectra of goat milk were used. The method was tested on 2 UV instruments. The comparison resulted in a significantly more robust (i.e., better) transferability between UV instruments for the partial least squares regression method on full spectra compared with previous univariate calibration of 4D spectra. Local (1 instrument) calibrations gave similar, significantly not different (chi-squared test) cross-validated prediction error results for the 2 methods. It can be concluded that there is no need for fourth derivation. Partial least squares regression on full spectra was equal or superior to using the 4D spectra. PMID:20630209

  12. New methods for the calibration of colour indices and O4 absorptions obtained from MAX-DOAS measurements

    NASA Astrophysics Data System (ADS)

    Wagner, Thomas; Beirle, Steffen; Dörner, Steffen; Remmers, Julia; Shaiganfar, Reza; Wang, Yang

    2015-04-01

    So called colour indices (CI) are defined as ratio of radiances at different wavelengths. CI measurements from Multi-AXis-Differential Absorption Spectroscopy (MAX-DOAS) observations are important for the detection and classification of clouds and aerosols. However, usually, MAX-DOAS instruments are not radiometrically calibrated. Thus the measured CI can not be directly compared to results from radiative transfer simulations. This shortcoming prevents the standardisation of cloud classification schemes. In addition to the CI, also the absorption of the oxygen dimer (O4) can be used for cloud detection, in particular for the detection of optically thick clouds or fog. However, also the retrieved O4 absorption has first to be calibrated (the absorption of the Fraunhofer reference spectrum has to be determined) before it can be used in a standardised way. We developed methods for the calibration of the CI and the O4 absorption derived from MAX-DOAS observations. They are based on the comparison of measurements and simulation results for well-defined atmospheric conditions. We estimate the accuracy of our calibration methods to < 10%.

  13. Reconciling Empirical Carbonate Clumped Isotope Calibrations: A Comparison of Calcite Precipitation and Acid Digestion Methods

    NASA Astrophysics Data System (ADS)

    Kelson, J.; Huntington, K. W.; Schauer, A. J.; Saenger, C.; Lechler, A. R.

    2015-12-01

    An accurate empirical calibration is necessary to confidently apply the carbonate clumped isotope (Δ47) thermometer. Previous synthetic carbonate calibrations disagree in temperature sensitivity, with one group of calibrations displaying a shallow Δ47-temperature slope (e.g., Dennis & Schrag, GCA, 2010), and the other a steep slope (e.g., Zaarur et al., EPSL, 2013). These calibrations differ in both the method of mineral precipitation and the temperature of the phosphoric acid used to digest carbonates for analysis, making it difficult to isolate the cause of the discrepancy. Here, we precipitate synthetic carbonates at temperatures of 6-80ºC using 4 different precipitation methods, and analyze the samples using both 90 and 25°C acid digestion. Precipitation experiments varied the use of salts (NaHCO3 and CaCl2) vs. dissolved CaCO3 as a starting solution, the use of carbonic anhydrase to promote isotopic equilibrium among dissolved inorganic carbon species in solution, and the method by which CO2 degasses to force carbonate precipitation. Carbonates precipitated by using salts and allowing CO2 to passively degas produce a shallow calibration slope that we hypothesize to approach isotopic equilibrium. Precipitation methods that bubble CO2 into solution then degas that CO2 (either passively or actively by bubbling N2) produce carbonates with consistently lower Δ47 and higher δ18O values for a given growth temperature. We infer that these carbonates grew in disequilibrium during rapid CO2 degassing. Varying acid digestion temperature does not change the results; acid fractionation factor is not correlated with grain size, Δ47, or d47 values. No precipitation method produces a steep calibration slope. Our large sample set of >60 carbonates lend confidence to a shallow slope calibration, and inform interpretations of Δ47 and δ18O values of natural carbonates that grow under conditions of isotopic disequilibrium.

  14. Method of rotation angle measurement in machine vision based on calibration pattern with spot array

    SciTech Connect

    Li Weimin; Jin Jing; Li Xiaofeng; Li Bin

    2010-02-20

    We propose a method of rotation angle measurement with high precision in machine vision. An area scan CCD camera, imaging lens, and calibration pattern with a spot array make up the measurement device for measuring the rotation angle. The calibration pattern with a spot array is installed at the rotation part, and the CCD camera is set at a certain distance from the rotation components. The coordinates of the spots on the calibration pattern is acquired through the vision image of the calibration pattern captured by the CCD camera. At the initial position of the calibration pattern, the camera is calibrated with the spot array; the mathematical model of distortion error of the CCD camera is built. With the equation of coordinate rotation measurement, the rotation angle of the spot array is detected. In the theoretic simulation, noise of different levels is added to the coordinates of the spot array. The experiment results show that the measurement device can measure the rotation angle precisely with a noncontact method. The standard deviation of rotation angle measurement is smaller than 3 arc sec. The measurement device can measure both microangles and large angles.

  15. Method of rotation angle measurement in machine vision based on calibration pattern with spot array.

    PubMed

    Li, Weimin; Jin, Jing; Li, Xiaofeng; Li, Bin

    2010-02-20

    We propose a method of rotation angle measurement with high precision in machine vision. An area scan CCD camera, imaging lens, and calibration pattern with a spot array make up the measurement device for measuring the rotation angle. The calibration pattern with a spot array is installed at the rotation part, and the CCD camera is set at a certain distance from the rotation components. The coordinates of the spots on the calibration pattern is acquired through the vision image of the calibration pattern captured by the CCD camera. At the initial position of the calibration pattern, the camera is calibrated with the spot array; the mathematical model of distortion error of the CCD camera is built. With the equation of coordinate rotation measurement, the rotation angle of the spot array is detected. In the theoretic simulation, noise of different levels is added to the coordinates of the spot array. The experiment results show that the measurement device can measure the rotation angle precisely with a noncontact method. The standard deviation of rotation angle measurement is smaller than 3 arc sec. The measurement device can measure both microangles and large angles. PMID:20174168

  16. An efficient calibration method for SQUID measurement system using three orthogonal Helmholtz coils

    NASA Astrophysics Data System (ADS)

    Hua, Li; Shu-Lin, Zhang; Chao-Xiang, Zhang; Xiang-Yan, Kong; Xiao-Ming, Xie

    2016-06-01

    For a practical superconducting quantum interference device (SQUID) based measurement system, the Tesla/volt coefficient must be accurately calibrated. In this paper, we propose a highly efficient method of calibrating a SQUID magnetometer system using three orthogonal Helmholtz coils. The Tesla/volt coefficient is regarded as the magnitude of a vector pointing to the normal direction of the pickup coil. By applying magnetic fields through a three-dimensional Helmholtz coil, the Tesla/volt coefficient can be directly calculated from magnetometer responses to the three orthogonally applied magnetic fields. Calibration with alternating current (AC) field is normally used for better signal-to-noise ratio in noisy urban environments and the results are compared with the direct current (DC) calibration to avoid possible effects due to eddy current. In our experiment, a calibration relative error of about 6.89 × 10‑4 is obtained, and the error is mainly caused by the non-orthogonality of three axes of the Helmholtz coils. The method does not need precise alignment of the magnetometer inside the Helmholtz coil. It can be used for the multichannel magnetometer system calibration effectively and accurately. Project supported by the “Strategic Priority Research Program (B)” of the Chinese Academy of Sciences (Grant No. XDB04020200) and the Shanghai Municipal Science and Technology Commission Project, China (Grant No. 15DZ1940902).

  17. Crystal timing offset calibration method for time of flight PET scanners

    NASA Astrophysics Data System (ADS)

    Ye, Jinghan; Song, Xiyun

    2016-03-01

    In time-of-flight (TOF) positron emission tomography (PET), precise calibration of the timing offset of each crystal of a PET scanner is essential. Conventionally this calibration requires a specially designed tool just for this purpose. In this study a method that uses a planar source to measure the crystal timing offsets (CTO) is developed. The method uses list mode acquisitions of a planar source placed at multiple orientations inside the PET scanner field-of-view (FOV). The placement of the planar source in each acquisition is automatically figured out from the measured data, so that a fixture for exactly placing the source is not required. The expected coincidence time difference for each detected list mode event can be found from the planar source placement and the detector geometry. A deviation of the measured time difference from the expected one is due to CTO of the two crystals. The least squared solution of the CTO is found iteratively using the list mode events. The effectiveness of the crystal timing calibration method is evidenced using phantom images generated by placing back each list mode event into the image space with the timing offset applied to each event. The zigzagged outlines of the phantoms in the images become smooth after the crystal timing calibration is applied. In conclusion, a crystal timing calibration method is developed. The method uses multiple list mode acquisitions of a planar source to find the least squared solution of crystal timing offsets.

  18. Evaluation of Linking Methods for Multidimensional IRT Calibrations

    ERIC Educational Resources Information Center

    Min, Kyung-Seok

    2007-01-01

    Most researchers agree that psychological/educational tests are sensitive to multiple traits, implying the need for a multidimensional item response theory (MIRT). One limitation of applying a MIRT in practice is the difficulty in establishing equivalent scales of multiple traits. In this study, a new MIRT linking method was proposed and evaluated…

  19. A GPS-Based Pitot-Static Calibration Method Using Global Output-Error Optimization

    NASA Technical Reports Server (NTRS)

    Foster, John V.; Cunningham, Kevin

    2010-01-01

    Pressure-based airspeed and altitude measurements for aircraft typically require calibration of the installed system to account for pressure sensing errors such as those due to local flow field effects. In some cases, calibration is used to meet requirements such as those specified in Federal Aviation Regulation Part 25. Several methods are used for in-flight pitot-static calibration including tower fly-by, pacer aircraft, and trailing cone methods. In the 1990 s, the introduction of satellite-based positioning systems to the civilian market enabled new inflight calibration methods based on accurate ground speed measurements provided by Global Positioning Systems (GPS). Use of GPS for airspeed calibration has many advantages such as accuracy, ease of portability (e.g. hand-held) and the flexibility of operating in airspace without the limitations of test range boundaries or ground telemetry support. The current research was motivated by the need for a rapid and statistically accurate method for in-flight calibration of pitot-static systems for remotely piloted, dynamically-scaled research aircraft. Current calibration methods were deemed not practical for this application because of confined test range size and limited flight time available for each sortie. A method was developed that uses high data rate measurements of static and total pressure, and GPSbased ground speed measurements to compute the pressure errors over a range of airspeed. The novel application of this approach is the use of system identification methods that rapidly compute optimal pressure error models with defined confidence intervals in nearreal time. This method has been demonstrated in flight tests and has shown 2- bounds of approximately 0.2 kts with an order of magnitude reduction in test time over other methods. As part of this experiment, a unique database of wind measurements was acquired concurrently with the flight experiments, for the purpose of experimental validation of the

  20. Research on calibration method of axis-shift multi-camera for aerial photogrammetry

    NASA Astrophysics Data System (ADS)

    Wang, Xiao; Fang, Junyong; Ma, Jingyu; Zhang, Xiaohong; Zhao, Dong; Liu, Xue

    2015-12-01

    Axis-shift multi-camera has been gradually applied in the aerial photogrammetry because of its advantages on structure design. In this paper, the basic axis-shift theory is analyzed, and an improved calibration method is described. A prototype system, including two axis-shift cameras, is developed to validate the feasibility and correctness of the proposed method. With the help of a high-precision indoor control field, the parameters of single camera and the relative orientation parameters of the dual camera system are calculated respectively. Experiment result indicates that this calibration method is suitable for the axis-shift multi camera system.

  1. Application of rational functions for the standard addition method.

    PubMed

    Gorazda, Katarzyna; Michałowska-Kaczmarczyk, Anna M; Asuero, Agustin G; Michałowski, Tadeusz

    2013-11-15

    Some rational functions are considered as the basis for calculation of unknown concentration (x0) of an analyte X determined according to the standard addition method (SAM). The correction for dilution of the sample tested during addition of successive increments of standard(ised) solution of X is involved in the related algorithm applied for calculation of the x0 value. The formulae derived were put in context with experimental data, obtained according to the AAS method from Cu-measurements in samples obtained by digestion of an ash obtained from incinerated sludge. It was stated that the use of rational functions for modeling purposes strengthens the robustness of the results thus obtained. PMID:24148496

  2. Calibrated propensity score method for survey nonresponse in cluster sampling

    PubMed Central

    Kim, Jae Kwang; Kwon, Yongchan; Paik, Myunghee Cho

    2016-01-01

    Weighting adjustment is commonly used in survey sampling to correct for unit nonresponse. In cluster sampling, the missingness indicators are often correlated within clusters and the response mechanism is subject to cluster-specific nonignorable missingness. Based on a parametric working model for the response mechanism that incorporates cluster-specific nonignorable missingness, we propose a method of weighting adjustment. We provide a consistent estimator of the mean or totals in cases where the study variable follows a generalized linear mixed-effects model. The proposed method is robust in the sense that the consistency of the estimator does not require correct specification of the functional forms of the response and outcome models. A consistent variance estimator based on Taylor linearization is also proposed. Numerical results, including a simulation and a real-data application, are presented.

  3. A robust method for online stereo camera self-calibration in unmanned vehicle system

    NASA Astrophysics Data System (ADS)

    Zhao, Yu; Chihara, Nobuhiro; Guo, Tao; Kimura, Nobutaka

    2014-06-01

    Self-calibration is a fundamental technology used to estimate the relative posture of the cameras for environment recognition in unmanned system. We focused on the issue of recognition accuracy decrease caused by the vibration of platform and conducted this research to achieve on-line self-calibration using feature point's registration and robust estimation of fundamental matrix. Three key factors in this respect are needed to be improved. Firstly, the feature mismatching exists resulting in the decrease of estimation accuracy of relative posture. The second, the conventional estimation method cannot satisfy both the estimation speed and calibration accuracy at the same tame. The third, some system intrinsic noises also lead greatly to the deviation of estimation results. In order to improve the calibration accuracy, estimation speed and system robustness for the practical implementation, we discuss and analyze the algorithms to make improvements on the stereo camera system to achieve on-line self-calibration. Based on the epipolar geometry and 3D images parallax, two geometry constraints are proposed to make the corresponding feature points search performed in a small search-range resulting in the improvement of matching accuracy and searching speed. Then, two conventional estimation algorithms are analyzed and evaluated for estimation accuracy and robustness. The third, Rigorous posture calculation method is proposed with consideration of the relative posture deviation of each separated parts in the stereo camera system. Validation experiments were performed with the stereo camera mounted on the Pen-Tilt Unit for accurate rotation control and the evaluation shows that our proposed method is fast and of high accuracy with high robustness for on-line self-calibration algorithm. Thus, as the main contribution, we proposed methods to solve the on-line self-calibration fast and accurately, envision the possibility for practical implementation on unmanned system as

  4. A new time calibration method for switched-capacitor-array-based waveform samplers

    SciTech Connect

    Kim, H.; Chen, C. -T.; Eclov, N.; Ronzhin, A.; Murat, P.; Ramberg, E.; Los, S.; Moses, W.; Choong, W. -S.; Kao, C. -M.

    2014-08-24

    Here we have developed a new time calibration method for the DRS4 waveform sampler that enables us to precisely measure the non-uniform sampling interval inherent in the switched-capacitor cells of the DRS4. The method uses the proportionality between the differential amplitude and sampling interval of adjacent switched-capacitor cells responding to a sawtooth-shape pulse. In the experiment, a sawtooth-shape pulse with a 40 ns period generated by a Tektronix AWG7102 is fed to a DRS4 evaluation board for calibrating the sampling intervals of all 1024 cells individually. The electronic time resolution of the DRS4 evaluation board with the new time calibration is measured to be ~2.4 ps RMS by using two simultaneous Gaussian pulses with 2.35 ns full-width at half-maximum and applying a Gaussian fit. The time resolution dependencies on the time difference with the new time calibration are measured and compared to results obtained by another method. Ultimately, the new method could be applicable for other switched-capacitor-array technology-based waveform samplers for precise time calibration.

  5. Enhanced Calibration Method of Silver Detector to Use in a Plasma Focus Device of Low Yield

    SciTech Connect

    Moreno, Jose; Soto, Leopoldo; Tarifeno-Saldivia, Ariel

    2010-08-04

    In this work, a calibration method based only in the use of a continuous neutron source as reference is presented. After the silver foil reached activation steady state, the neutron source was removed. The accumulated were analyzed by the method of linear least squares.

  6. Two-dimensional differential calibration method for a neutron dosemeter using a thermal neutron beam.

    PubMed

    Matsumoto, Tetsuro; Harano, Hideki; Masuda, Akihiko; Nishiyama, Jun; Matsue, Hideaki; Uritani, Akira; Nunomiya, Tomoya

    2013-08-01

    A new thermal neutron calibration method to experimentally determine the energy response function of a neutron detector using a pulse parallel beam and the time-of-flight (TOF) technique is developed. The calibration method was experimentally demonstrated for a (3)He proportional counter and an electric personal dosemeter using a pulsed thermal neutron beam from the research reactor JRR-3M. The responses of the detectors were successfully obtained as a function of neutron energy. However, detailed information on the detector structure is required to obtain the spatial response distribution for the detector. The authors further propose an improved calibration method obtaining the spatial response distribution using a pulsed narrow beam, the TOF technique and a beam scanning technique. PMID:23509397

  7. A simple method for wind tunnel balance calibration including non-linear interaction terms

    NASA Astrophysics Data System (ADS)

    Ramaswamy, M. A.; Srinivas, T.; Holla, V. S.

    The conventional method for calibrating wind tunnel balances to obtain the coupled linear and nonlinear interaction terms requires the application of combinations of pure components of the loads on the calibration body compensating the deflection of the balance. For a six-component balance, this calls for a complex loading system and an arrangement to translate and tilt the balance support about all three axes. A simple method called the least-square method is illustrated for a three-component balance. The simplicity arises from the fact that application of the pure components of the loads or reorientation of the balance is not required. A single load is applied that has various components whose magnitudes can be easily found knowing the orientation of the calibration body under load and the point of application of the load. The coefficients are obtained by using the least-square-fit approach to match the outputs obtained for various combinations of load.

  8. Optical System Error Analysis and Calibration Method of High-Accuracy Star Trackers

    PubMed Central

    Sun, Ting; Xing, Fei; You, Zheng

    2013-01-01

    The star tracker is a high-accuracy attitude measurement device widely used in spacecraft. Its performance depends largely on the precision of the optical system parameters. Therefore, the analysis of the optical system parameter errors and a precise calibration model are crucial to the accuracy of the star tracker. Research in this field is relatively lacking a systematic and universal analysis up to now. This paper proposes in detail an approach for the synthetic error analysis of the star tracker, without the complicated theoretical derivation. This approach can determine the error propagation relationship of the star tracker, and can build intuitively and systematically an error model. The analysis results can be used as a foundation and a guide for the optical design, calibration, and compensation of the star tracker. A calibration experiment is designed and conducted. Excellent calibration results are achieved based on the calibration model. To summarize, the error analysis approach and the calibration method are proved to be adequate and precise, and could provide an important guarantee for the design, manufacture, and measurement of high-accuracy star trackers. PMID:23567527

  9. Optical system error analysis and calibration method of high-accuracy star trackers.

    PubMed

    Sun, Ting; Xing, Fei; You, Zheng

    2013-01-01

    The star tracker is a high-accuracy attitude measurement device widely used in spacecraft. Its performance depends largely on the precision of the optical system parameters. Therefore, the analysis of the optical system parameter errors and a precise calibration model are crucial to the accuracy of the star tracker. Research in this field is relatively lacking a systematic and universal analysis up to now. This paper proposes in detail an approach for the synthetic error analysis of the star tracker, without the complicated theoretical derivation. This approach can determine the error propagation relationship of the star tracker, and can build intuitively and systematically an error model. The analysis results can be used as a foundation and a guide for the optical design, calibration, and compensation of the star tracker. A calibration experiment is designed and conducted. Excellent calibration results are achieved based on the calibration model. To summarize, the error analysis approach and the calibration method are proved to be adequate and precise, and could provide an important guarantee for the design, manufacture, and measurement of high-accuracy star trackers. PMID:23567527

  10. Standardization of Laser Methods and Techniques for Vibration Measurements and Calibrations

    SciTech Connect

    Martens, Hans-Juergen von

    2010-05-28

    The realization and dissemination of the SI units of motion quantities (vibration and shock) have been based on laser interferometer methods specified in international documentary standards. New and refined laser methods and techniques developed by national metrology institutes and by leading manufacturers in the past two decades have been swiftly specified as standard methods for inclusion into in the series ISO 16063 of international documentary standards. A survey of ISO Standards for the calibration of vibration and shock transducers demonstrates the extended ranges and improved accuracy (measurement uncertainty) of laser methods and techniques for vibration and shock measurements and calibrations. The first standard for the calibration of laser vibrometers by laser interferometry or by a reference accelerometer calibrated by laser interferometry (ISO 16063-41) is on the stage of a Draft International Standard (DIS) and may be issued by the end of 2010. The standard methods with refined techniques proved to achieve wider measurement ranges and smaller measurement uncertainties than that specified in the ISO Standards. The applicability of different standardized interferometer methods to vibrations at high frequencies was recently demonstrated up to 347 kHz (acceleration amplitudes up to 350 km/s{sup 2}). The relative deviations between the amplitude measurement results of the different interferometer methods that were applied simultaneously, differed by less than 1% in all cases.

  11. Application of the MCMC Method for the Calibration of DSMC Parameters

    SciTech Connect

    Strand, James S.; Goldstein, David B.

    2011-05-20

    A Markov Chain Monte Carlo (MCMC) algorithm was employed to obtain a calibrated distribution for the hard sphere diameter, the VHS reference diameter, and the temperature viscosity exponent of argon, for use in the Direct Simulation Monte Carlo (DSMC) method. Shock-tube experiments from Alsmeyer [1] were used to provide the necessary calibration data for use in the MCMC method. The DSMC method used in this work employs the algorithm of Bird [2], with modifications to allow for the efficient simulation of a 1D shock. When calibrating for the hard sphere diameter (the temperature viscosity exponent is set to 0.5 for the hard sphere method), the results of the MCMC method agree with a simple brute-force method, and a single value for the hard-sphere diameter is obtained. For the VHS method, however, when simultaneously calibrating the VHS reference diameter and the temperature viscosity exponent, we find that normalized density data alone does not provide sufficient information to obtain a single solution for both parameters. Instead we find a band in parameter space where acceptable solutions are obtained.

  12. Calibration method for a vision guiding-based laser-tracking measurement system

    NASA Astrophysics Data System (ADS)

    Shao, Mingwei; Wei, Zhenzhong; Hu, Mengjie; Zhang, Guangjun

    2015-08-01

    Laser-tracking measurement systems (laser trackers) based on a vision-guiding device are widely used in industrial fields, and their calibration is important. As conventional methods typically have many disadvantages, such as difficult machining of the target and overdependence on the retroreflector, a novel calibration method is presented in this paper. The retroreflector, which is necessary in the normal calibration method, is unnecessary in our approach. As the laser beam is linear, points on the beam can be obtained with the help of a normal planar target. In this way, we can determine the function of a laser beam under the camera coordinate system, while its corresponding function under the laser-tracker coordinate system can be obtained from the encoder of the laser tracker. Clearly, when several groups of functions are confirmed, the rotation matrix can be solved from the direction vectors of the laser beams in different coordinate systems. As the intersection of the laser beams is the origin of the laser-tracker coordinate system, the translation matrix can also be determined. Our proposed method not only achieves the calibration of a single laser-tracking measurement system but also provides a reference for the calibration of a multistation system. Simulations to evaluate the effects of some critical factors were conducted. These simulations show the robustness and accuracy of our method. In real experiments, the root mean square error of the calibration result reached 1.46 mm within a range of 10 m, even though the vision-guiding device focuses on a point approximately 5 m away from the origin of its coordinate system, with a field of view of approximately 200 mm  ×  200 mm.

  13. A geometric calibration method for inverse geometry computed tomography using P-matrices

    NASA Astrophysics Data System (ADS)

    Slagowski, Jordan M.; Dunkerley, David A. P.; Hatt, Charles R.; Speidel, Michael A.

    2016-03-01

    Accurate and artifact free reconstruction of tomographic images requires precise knowledge of the imaging system geometry. This work proposes a novel projection matrix (P-matrix) based calibration method to enable C-arm inverse geometry CT (IGCT). The method is evaluated for scanning-beam digital x-ray (SBDX), a C-arm mounted inverse geometry fluoroscopic technology. A helical configuration of fiducials is imaged at each gantry angle in a rotational acquisition. For each gantry angle, digital tomosynthesis is performed at multiple planes and a composite image analogous to a cone-beam projection is generated from the plane stack. The geometry of the C-arm, source array, and detector array is determined at each angle by constructing a parameterized 3D-to-2D projection matrix that minimizes the sum-of-squared deviations between measured and projected fiducial coordinates. Simulations were used to evaluate calibration performance with translations and rotations of the source and detector. In a geometry with 1 mm translation of the central ray relative to the axis-of-rotation and 1 degree yaw of the detector and source arrays, the maximum error in the recovered translational parameters was 0.4 mm and maximum error in the rotation parameter was 0.02 degrees. The relative rootmean- square error in a reconstruction of a numerical thorax phantom was 0.4% using the calibration method, versus 7.7% without calibration. Changes in source-detector-distance were the most challenging to estimate. Reconstruction of experimental SBDX data using the proposed method eliminated double contour artifacts present in a non-calibrated reconstruction. The proposed IGCT geometric calibration method reduces image artifacts when uncertainties exist in system geometry.

  14. CEILINEX 2015: Validation of calibration methods during the ceilometer inter-comparison

    NASA Astrophysics Data System (ADS)

    Hervo, Maxime

    2016-04-01

    In Europe, more than 700 ceilometers are measuring continuously. These instruments can be used for many applications such as detection of cloud base and aerosol layers height, aerosol profiling or for fog now-casting. However, from different manufacturers exist and the results can vary extensively from one type to another. During the CeiLinEx2015 campaign (Ceilometer Performance Experiment at Lindenberg 2015), 6 types of ceilometers (CL31, CL51, CHM15k, CHM15kx CS135 and LD40) were measuring simultaneously at Lindenberg (Germany) from June to September 2015. Each type was represented by two instruments in order to assess the instrument-to-instrument variability. A companion contribution by Mattis et al. presents an overview of the campaign. The monitoring of the temporal and spatial evolution of aerosol layers like the volcanic ash, is crucial to compare measurements from different sites. Therefore, all instruments need to be calibrated in order to provide consistent results. This contribution will focus on the validation and the comparison of state-of-the-art calibration methods. The calibration methods tested were are the cloud calibration (O'Connor et al., 2004) and the Rayleigh calibration Method (Wiegner and Geiß, 2012). Both methods can be applied without on-site intervention and are thus suitable for automatic networks. Operational automated algorithms based on these methods were developed in the framework of the TOPROF project (ESSEM COST Action ES1303)..The Cloud calibration was found more appropriate for analog instruments with analog signal detection measuring at around 905nm (Vaisala CL51 and CL31 and Campbell Scientific CS135). The Rayleigh was more suitable for photon-counting systems measuring at 1064nm (CHM15k, CHM15kx). For the first time, these methods were tested simultaneously on different instrument types and compared amongst each other. For a dust event occurred on the 13/08/2015, the attenuated backscatter coefficient difference amongst all

  15. Weighted partial least squares method to improve calibration precision for spectroscopic noise-limited data

    SciTech Connect

    Haaland, D.M.; Jones, H.D.T.

    1997-09-01

    Multivariate calibration methods have been applied extensively to the quantitative analysis of Fourier transform infrared (FT-IR) spectral data. Partial least squares (PLS) methods have become the most widely used multivariate method for quantitative spectroscopic analyses. Most often these methods are limited by model error or the accuracy or precision of the reference methods. However, in some cases, the precision of the quantitative analysis is limited by the noise in the spectroscopic signal. In these situations, the precision of the PLS calibrations and predictions can be improved by the incorporation of weighting in the PLS algorithm. If the spectral noise of the system is known (e.g., in the case of detector-noise-limited cases), then appropriate weighting can be incorporated into the multivariate spectral calibrations and predictions. A weighted PLS (WPLS) algorithm was developed to improve the precision of the analysis in the case of spectral-noise-limited data. This new PLS algorithm was then tested with real and simulated data, and the results compared with the unweighted PLS algorithm. Using near-infrared (NIR) calibration precision when the WPLS algorithm was applied. The best WPLS method improved prediction precision for the analysis of one of the minor components by a factor of nearly 9 relative to the unweighted PLS algorithm.

  16. A Novel Nonintrusive Method to Resolve the Thermal Dome Effect of Pyranometers: Radiometric Calibration and Implications

    NASA Technical Reports Server (NTRS)

    Ji. Q.; Tsay, S.-C.; Lau, K. M.; Hansell, R. A.; Butler, J. J.; Cooper, J. W.

    2011-01-01

    Traditionally the calibration equation for pyranometers assumes that the measured solar irradiance is solely proportional to the thermopile s output voltage; therefore, only a single calibration factor is derived. This causes additional measurement uncertainties because it does not capture sufficient information to correctly account for a pyranometer s thermal effect. In our updated calibration equation, temperatures from the pyranometer's dome and case are incorporated to describe the instrument's thermal behavior, and a new set of calibration constants are determined, thereby reducing measurement uncertainties. In this paper, we demonstrate why a pyranometer's uncertainty using the traditional calibration equation is always larger than a few percent, but with the new approach can become much less than 1% after the thermal issue is resolved. The highlighted calibration results are based on NIST traceable light sources under controlled laboratory conditions. The significance of the new approach lends itself to not only avoiding the uncertainty caused by a pyranometer's thermal effect but also the opportunity to better isolate and characterize other instrumental artifacts, such as angular response and nonlinearity of the thermopile, to further reduce additional uncertainties. We also discuss some of the implications, including an example of how the thermal issue can potentially impact climate studies by evaluating aerosol s direct radiative effect using field measurements with and without considering the pyranometer s thermal effect. The results of radiative transfer model simulation show that a pyranometer s thermal effect on solar irradiance measurements at the surface can be translated into a significant alteration of the calculated distribution of solar energy inside the column atmosphere.

  17. An improved method for calibrating the gantry angles of linear accelerators.

    PubMed

    Higgins, Kyle; Treas, Jared; Jones, Andrew; Fallahian, Naz Afarin; Simpson, David

    2013-11-01

    Linear particle accelerators (linacs) are widely used in radiotherapy procedures; therefore, accurate calibrations of gantry angles must be performed to prevent the exposure of healthy tissue to excessive radiation. One of the common methods for calibrating these angles is the spirit level method. In this study, a new technique for calibrating the gantry angle of a linear accelerator was examined. A cubic phantom was constructed of Styrofoam with small lead balls, embedded at specific locations in this foam block. Several x-ray images were taken of this phantom at various gantry angles using an electronic portal imaging device on the linac. The deviation of the gantry angles were determined by analyzing the images using a customized computer program written in ImageJ (National Institutes of Health). Gantry angles of 0, 90, 180, and 270 degrees were chosen and the results of both calibration methods were compared for each of these angles. The results revealed that the image method was more precise than the spirit level method. For the image method, the average of the measured values for the selected angles of 0, 90, 180, and 270 degrees were found to be -0.086 ± 0.011, 90.018 ± 0.011, 180.178 ± 0.015, and 269.972 ± 0.006 degrees, respectively. The corresponding average values using the spirit level method were 0.2 ± 0.03, 90.2 ± 0.04, 180.1 ± 0.01, and 269.9 ± 0.05 degrees, respectively. Based on these findings, the new method was shown to be a reliable technique for calibrating the gantry angle. PMID:24077078

  18. Sensitivity analysis of a geometric calibration method using projection matrices for digital tomosynthesis systems

    SciTech Connect

    Li Xinhua; Zhang Da; Liu, Bob

    2011-01-15

    Purpose: To study the sensitivity of a geometric calibration method using projection matrices for digital tomosynthesis systems. Methods: A generic geometric calibration method for tomographic imaging systems has been presented in our previous work. The method involves a scan of a calibration phantom with multiple markers. Their locations in projection images are detected and are associated with their 3D coordinates to compute 3x4 projection matrices, which can be used in subsequent image reconstruction. The accuracy of geometric calibration may be affected by errors in the input data of marker positions. The effects of errors may depend on the number of markers and the volume surrounded by them in 3D space. This work analyzed the sensitivity of the calibration method to the above factors. A 6 cm CIRS breast research phantom and a prototype breast tomosynthesis system were used for our tests. A high contrast ring and two small speck groups were reconstructed in various testing cases for comparison. To achieve quantitative assessment, a 15x15 point detection mask was adopted for detecting signals and for computing changes between testing cases and the regular geometric calibration. Results: When 3D coordinates and 2D projections of markers were accurate, all tested numbers of markers, 6-44, provided similar high quality reconstructions of the ring and the two speck groups. Errors in marker positions resulted in image degradations and signal changes, which increased with fewer markers and smaller volume surrounded by markers in the 3D object space. Signal changes of small specks were more significant than those of the ring. Errors in marker projections produced drastic image degradations. Coplanar marker placement caused a failure in projection matrix computation. Conclusions: For practical geometric calibration phantom design, ample markers are desired. They need to have a large volumetric coverage in the 3D space and be far from being coplanar. Precise

  19. Calibrating CAT Pools and Online Pretest Items Using Marginal Maximum Likelihood Methods.

    ERIC Educational Resources Information Center

    Pommerich, Mary; Segall, Daniel O.

    Research discussed in this paper was conducted as part of an ongoing large-scale simulation study to evaluate methods of calibrating pretest items for computerized adaptive testing (CAT) pools. The simulation was designed to mimic the operational CAT Armed Services Vocational Aptitude Battery (ASVAB) testing program, in which a single pretest item…

  20. Calibration method for line-structured light vision sensor based on a single ball target

    NASA Astrophysics Data System (ADS)

    Liu, Zhen; Li, Xiaojing; Li, Fengjiao; Zhang, Guangjun

    2015-06-01

    Profile feature imaging for ball targets is unaffected by the position of the target. On this basis, this study proposes a method for the rapid calibration of a line-structured light system based on a single ball target. The calibration process is as follows: the ball target is placed at least once and is illuminated by the light stripe from the laser projector. The vision sensor captures an image of this target. The laser stripe and profile images of the ball target are then extracted. Based on these extracted features and the optical centre of the camera, the spatial equations of the ball target and a cone profile are calculated. The plane on which the intersection line of the two equations lies is the light plane. Finally, the optimal solution for the light plane equation is obtained through nonlinear optimization under a maximum likelihood criterion. The validity of the proposed method is demonstrated through simulation and physical experiments. In the physical experiment, the field of view of the structured light vision sensor measures 300 mm×250 mm. A calibration accuracy of 0.04 mm can be achieved using the proposed method. This accuracy is comparable to that of the calibration method which utilizes planar targets.

  1. A method for atomic force microscopy cantilever stiffness calibration under heavy fluid loading

    SciTech Connect

    Kennedy, Scott J.; Cole, Daniel G.; Clark, Robert L.

    2009-12-15

    This work presents a method for force calibration of rectangular atomic force microscopy (AFM) microcantilevers under heavy fluid loading. Theoretical modeling of the thermal response of microcantilevers is discussed including a fluid-structure interaction model of the cantilever-fluid system that incorporates the results of the fluctuation-dissipation theorem. This model is curve fit to the measured thermal response of a cantilever in de-ionized water and a cost function is used to quantify the difference between the theoretical model and measured data. The curve fit is performed in a way that restricts the search space to parameters that reflect heavy fluid loading conditions. The resulting fitting parameters are used to calibrate the cantilever. For comparison, cantilevers are calibrated using Sader's method in air and the thermal noise method in both air and water. For a set of eight cantilevers ranging in stiffness from 0.050 to 5.8 N/m, the maximum difference between Sader's calibration performed in air and the new method performed in water was 9.4%. A set of three cantilevers that violate the aspect ratio assumption associated with the fluid loading model (length-to-width ratios less than 3.5) ranged in stiffness from 0.85 to 4.7 N/m and yielded differences as high as 17.8%.

  2. Comparison of "E-Rater"[R] Automated Essay Scoring Model Calibration Methods Based on Distributional Targets

    ERIC Educational Resources Information Center

    Zhang, Mo; Williamson, David M.; Breyer, F. Jay; Trapani, Catherine

    2012-01-01

    This article describes two separate, related studies that provide insight into the effectiveness of "e-rater" score calibration methods based on different distributional targets. In the first study, we developed and evaluated a new type of "e-rater" scoring model that was cost-effective and applicable under conditions of absent human rating and…

  3. A new time calibration method for switched-capacitor-array-based waveform samplers

    DOE PAGESBeta

    Kim, H.; Chen, C. -T.; Eclov, N.; Ronzhin, A.; Murat, P.; Ramberg, E.; Los, S.; Moses, W.; Choong, W. -S.; Kao, C. -M.

    2014-08-24

    Here we have developed a new time calibration method for the DRS4 waveform sampler that enables us to precisely measure the non-uniform sampling interval inherent in the switched-capacitor cells of the DRS4. The method uses the proportionality between the differential amplitude and sampling interval of adjacent switched-capacitor cells responding to a sawtooth-shape pulse. In the experiment, a sawtooth-shape pulse with a 40 ns period generated by a Tektronix AWG7102 is fed to a DRS4 evaluation board for calibrating the sampling intervals of all 1024 cells individually. The electronic time resolution of the DRS4 evaluation board with the new time calibrationmore » is measured to be ~2.4 ps RMS by using two simultaneous Gaussian pulses with 2.35 ns full-width at half-maximum and applying a Gaussian fit. The time resolution dependencies on the time difference with the new time calibration are measured and compared to results obtained by another method. Ultimately, the new method could be applicable for other switched-capacitor-array technology-based waveform samplers for precise time calibration.« less

  4. Multi-point radiometric calibration method based on complex spectrum of Fourier transform spectrometer

    NASA Astrophysics Data System (ADS)

    He, Qian; Wang, Guangping; Wu, Jingli; Li, Junwei

    2014-11-01

    As the impact of the instrument internal error, external interference and other factors, the interferogram measured by Fourier transform spectrometer is asymmetric, result in the complex outcome after Fourier transform. Currently, most radiometric calibration method used for Fourier transform spectrometer is usually based on real spectrums, which is converted from the above complex spectrum by calculating magnitude value or make the phase correction first. Proceed from error sources and mechanisms of the Fourier transform spectrometer, we propose a multi-point radiometric calibration method based on complex spectral data to improve the processing efficiency and accuracy, which is obtained by the original interferogram via Fourier transform. We solving the instrument response function include linear gain and offset by complex spectrum above to calculate complex spectral radiance. Compared with the traditional method based on real spectrum, the present efficient method does not limited to real spectrum and the phase correction is not required. In this paper, we use BOMEM's MR304 Fourier transform infrared spectrometer and the DCN1000N3 blackbody made by HGH Infrared Systems to carry out the radiation calibration experiment in laboratory. The results show that, the amplitude of complex radiance spectrum obtained by this method has a high consistency with the theoretical value, while the extra imaginary spectrum is similar with the difference between results and theoretical value in absolute value and trends. It proved that, this multi-point radiometric calibration method by using the amplitude of complex spectral data is highly reliable; meanwhile, the imaginary spectrum can reflect the calibration error very well and offer a new technical approach for accuracy evaluation research.

  5. Energy Calibration of the BaBar EMC Using the Pi0 Invariant Mass Method

    SciTech Connect

    Tanner, David J.; /Manchester U.

    2007-04-06

    The BaBar electromagnetic calorimeter energy calibration method was compared with the local and global peak iteration procedures, of Crystal Barrel and CLEO-II. An investigation was made of the possibility of {Upsilon}(4S) background reduction which could lead to increased statistics over a shorter time interval, for efficient calibration runs. The BaBar software package was used with unreconstructed data to study the energy response of the calorimeter, by utilizing the {pi}{sup 0} mass constraint on pairs of photon clusters.

  6. Method and system for calibrating acquired spectra for use in spectral analysis

    DOEpatents

    Reber, Edward L.; Rohde, Kenneth W.; Blackwood, Larry G.

    2010-09-14

    A method for calibrating acquired spectra for use in spectral analysis includes performing Gaussian peak fitting to spectra acquired by a plurality of NaI detectors to define peak regions. A Na and annihilation doublet may be located among the peak regions. A predetermined energy level may be applied to one of the peaks in the doublet and a location of a hydrogen peak may be predicted based on the location of at least one of the peaks of the doublet. Control systems for calibrating spectra are also disclosed.

  7. A curve fitting method for extrinsic camera calibration from a single image of a cylindrical object

    NASA Astrophysics Data System (ADS)

    Winkler, A. W.; Zagar, B. G.

    2013-08-01

    An important step in the process of optical steel coil quality assurance is to measure the proportions of width and radius of steel coils as well as the relative position and orientation of the camera. This work attempts to estimate these extrinsic parameters from single images by using the cylindrical coil itself as the calibration target. Therefore, an adaptive least-squares algorithm is applied to fit parametrized curves to the detected true coil outline in the acquisition. The employed model allows for strictly separating the intrinsic and the extrinsic parameters. Thus, the intrinsic camera parameters can be calibrated beforehand using available calibration software. Furthermore, a way to segment the true coil outline in the acquired images is motivated. The proposed optimization method yields highly accurate results and can be generalized even to measure other solids which cannot be characterized by the identification of simple geometric primitives.

  8. Cosmological models and gamma-ray bursts calibrated by using Padé method

    NASA Astrophysics Data System (ADS)

    Liu, Jing; Wei, Hao

    2015-11-01

    Gamma-ray bursts (GRBs) are among the most powerful sources in the universe. In the recent years, GRBs have been proposed as a complementary probe to type Ia supernovae. However, as is well known, there is a circularity problem in the use of GRBs to study cosmology. In this work, based on the Padé approximant, we propose a new cosmology-independent method to calibrate GRBs. We consider a sample consisting of 138 long Swift GRBs and obtain 79 calibrated long GRBs at high-redshift z>1.4 (named Mayflower sample) which can be used to constrain cosmological models without the circularity problem. Then, we consider the constraints on several cosmological models with these 79 calibrated GRBs and other observational data. We show that GRBs are competent to be a complementary probe to the other well-established cosmological observations.

  9. Investigation of infrared calibration methods for application to the study of methyl methacrylate polymerization.

    PubMed

    Kaczmarczyk, B; Morejko-Buz, B; Stolarzewicz, A

    2001-08-01

    Infrared spectroscopy has been used to monitor the polymerization of methyl methacrylate. Concentrations of methyl methacrylate in the reaction mixture were determined by use of three calibration methods. Classical quantitative analysis was used to measure the height of the stretching vibration bands of the vinyl group at 1639 cm(-1). A calibration procedure using the considerably higher intensity of the C = O stretching vibration band of the carbonyl ester group at 1725 cm(-1) seemed useful only for high concentrations of methyl methacrylate, i.e. at the beginning of reaction, because this band overlaps that of poly(methyl methacrylate). Use of second-derivative spectra and measuring their values at 1725 cm(-1) enabled estimation of ten times lower concentrations of methyl methacrylate the calibration using the band from the vinyl group. PMID:11569872

  10. Calibrating system errors of large scale three-dimensional profile measurement instruments by subaperture stitching method.

    PubMed

    Dong, Zhichao; Cheng, Haobo; Feng, Yunpeng; Su, Jingshi; Wu, Hengyu; Tam, Hon-Yuen

    2015-07-01

    This study presents a subaperture stitching method to calibrate system errors of several ∼2  m large scale 3D profile measurement instruments (PMIs). The calibration process was carried out by measuring a Φ460  mm standard flat sample multiple times at different sites of the PMI with a length gauge; then the subaperture data were stitched together using a sequential or simultaneous stitching algorithm that minimizes the inconsistency (i.e., difference) of the discrete data in the overlapped areas. The system error can be used to compensate the measurement results of not only large flats, but also spheres and aspheres. The feasibility of the calibration was validated by measuring a Φ1070  mm aspheric mirror, which can raise the measurement accuracy of PMIs and provide more reliable 3D surface profiles for guiding grinding, lapping, and even initial polishing processes. PMID:26193139

  11. Method and apparatus for calibrating multi-axis load cells in a dexterous robot

    NASA Technical Reports Server (NTRS)

    Wampler, II, Charles W. (Inventor); Platt, Jr., Robert J. (Inventor)

    2012-01-01

    A robotic system includes a dexterous robot having robotic joints, angle sensors adapted for measuring joint angles at a corresponding one of the joints, load cells for measuring a set of strain values imparted to a corresponding one of the load cells during a predetermined pose of the robot, and a host machine. The host machine is electrically connected to the load cells and angle sensors, and receives the joint angle values and strain values during the predetermined pose. The robot presses together mating pairs of load cells to form the poses. The host machine executes an algorithm to process the joint angles and strain values, and from the set of all calibration matrices that minimize error in force balance equations, selects the set of calibration matrices that is closest in a value to a pre-specified value. A method for calibrating the load cells via the algorithm is also provided.

  12. A novel implementation of homodyne time interval analysis method for primary vibration calibration

    NASA Astrophysics Data System (ADS)

    Sun, Qiao; Zhou, Ling; Cai, Chenguang; Hu, Hongbo

    2011-12-01

    In this paper, the shortcomings and their causes of the conventional homodyne time interval analysis (TIA) method is described with respect to its software algorithm and hardware implementation, based on which a simplified TIA method is proposed with the help of virtual instrument technology. Equipped with an ordinary Michelson interferometer and dual channel synchronous data acquisition card, the primary vibration calibration system using the simplified method can perform measurements of complex sensitivity of accelerometers accurately, meeting the uncertainty requirements laid down in pertaining ISO standard. The validity and accuracy of the simplified TIA method is verified by simulation and comparison experiments with its performance analyzed. This simplified method is recommended to apply in national metrology institute of developing countries and industrial primary vibration calibration labs for its simplified algorithm and low requirements on hardware.

  13. Assessment of measurement errors and dynamic calibration methods for three different tipping bucket rain gauges

    NASA Astrophysics Data System (ADS)

    Shedekar, Vinayak S.; King, Kevin W.; Fausey, Norman R.; Soboyejo, Alfred B. O.; Harmel, R. Daren; Brown, Larry C.

    2016-09-01

    Three different models of tipping bucket rain gauges (TBRs), viz. HS-TB3 (Hydrological Services Pty Ltd.), ISCO-674 (Isco, Inc.) and TR-525 (Texas Electronics, Inc.), were calibrated in the lab to quantify measurement errors across a range of rainfall intensities (5 mm·h- 1 to 250 mm·h- 1) and three different volumetric settings. Instantaneous and cumulative values of simulated rainfall were recorded at 1, 2, 5, 10 and 20-min intervals. All three TBR models showed a substantial deviation (α = 0.05) in measurements from actual rainfall depths, with increasing underestimation errors at greater rainfall intensities. Simple linear regression equations were developed for each TBR to correct the TBR readings based on measured intensities (R2 > 0.98). Additionally, two dynamic calibration techniques, viz. quadratic model (R2 > 0.7) and T vs. 1/Q model (R2 = > 0.98), were tested and found to be useful in situations when the volumetric settings of TBRs are unknown. The correction models were successfully applied to correct field-collected rainfall data from respective TBR models. The calibration parameters of correction models were found to be highly sensitive to changes in volumetric calibration of TBRs. Overall, the HS-TB3 model (with a better protected tipping bucket mechanism, and consistent measurement errors across a range of rainfall intensities) was found to be the most reliable and consistent for rainfall measurements, followed by the ISCO-674 (with susceptibility to clogging and relatively smaller measurement errors across a range of rainfall intensities) and the TR-525 (with high susceptibility to clogging and frequent changes in volumetric calibration, and highly intensity-dependent measurement errors). The study demonstrated that corrections based on dynamic and volumetric calibration can only help minimize-but not completely eliminate the measurement errors. The findings from this study will be useful for correcting field data from TBRs; and may have major

  14. A preliminary investigation of a method to calibrate strain gauge balances by means of a reference balance

    NASA Astrophysics Data System (ADS)

    Han, Buzhang; Wang, Mingying; Johnson, G. I.; Widmark, S.

    1992-06-01

    This investigation was started in order to obtain preliminary experience on the concept of a reference balance planned to be used in the T1500 calibration rig. The reference balance was simulated by using an available half model balance in the MK5 calibration rig at FFA. First the reference balance was calibrated and then the result was used for calibration of an ordinary six component sting balance attached with its model end to the half model balance. Both balances had been previously calibrated by the reposition method. Evaluation of the reference balance nonrepositioning concept was obtained by comparison of results.

  15. Optics-Only Calibration of a Neural-Net Based Optical NDE Method for Structural Health Monitoring

    NASA Technical Reports Server (NTRS)

    Decker, Arthur J.

    2004-01-01

    A calibration process is presented that uses optical measurements alone to calibrate a neural-net based NDE method. The method itself detects small changes in the vibration mode shapes of structures. The optics-only calibration process confirms previous work that the sensitivity to vibration-amplitude changes can be as small as 10 nanometers. A more practical value in an NDE service laboratory is shown to be 50 nanometers. Both model-generated and experimental calibrations are demonstrated using two implementations of the calibration technique. The implementations are based on previously published demonstrations of the NDE method and an alternative calibration procedure that depends on comparing neural-net and point sensor measurements. The optics-only calibration method, unlike the alternative method, does not require modifications of the structure being tested or the creation of calibration objects. The calibration process can be used to test improvements in the NDE process and to develop a vibration-mode-independence of damagedetection sensitivity. The calibration effort was intended to support NASA s objective to promote safety in the operations of ground test facilities or aviation safety, in general, by allowing the detection of the gradual onset of structural changes and damage.

  16. VIIRS Day-Night Band (DNB) calibration methods for improved uniformity

    NASA Astrophysics Data System (ADS)

    Mills, Stephen; Miller, Steven D.

    2014-10-01

    The Suomi-NPP VIIRS Day-Night Band (DNB) offers quantitative measurements of visible and near-infrared light over a dynamic range from full daylight to the dimmest nighttime scenes. This range presents a challenge to radiometric calibration, but the instrument has exceeded all of its absolute radiometric requirements. Nevertheless, striping and banding are still visible, day or night, but especially in low-light scenes. The causes may be cross talk, stray light or hysteresis in the data used for calibration. These issues combine to reduce the utility of these unique observations for gaining new insight on the nocturnal environment. This paper presents methods for improving gain and offset uniformity for both day and night scenes while maintaining absolute radiometric accuracy. We evaluate removal of fixed-pattern non-uniformity in dark scenes on a per orbit basis using three different techniques: i) tracking the darkest 25th percentile calibration sector signal; ii) taking the mean of filtered dark Earth-view scenes to determine offset; iii) minimizing correlated error for dark scenes within an aggregation zone. For gain uniformity we discuss some problems with the current calibration methods, and demonstrate a technique to minimize the correlated error between detectors and aggregation zones using the moment matching technique for moonlit scenes. A similar technique can be used for daytime and twilight scenes. An alternative cross-calibration technique between gain stages uses indirect illumination of solar diffuser view. The use of the space view and blackbody view for cross-calibration is also discussed. Histogram equalization is discussed for minimizing striping and banding. In all cases, data with stray light is filtered out to prevent contamination of the destriping process.

  17. Absolute calibration method for nanosecond-resolved, time-streaked, fiber optic light collection, spectroscopy systems

    SciTech Connect

    Johnston, Mark D.; Oliver, Bryan V.; Droemer, Darryl W.; Frogget, Brent; Crain, Marlon D.; Maron, Yitzhak

    2012-08-15

    This paper describes a convenient and accurate method to calibrate fast (<1 ns resolution) streaked, fiber optic light collection, spectroscopy systems. Such systems are inherently difficult to calibrate due to the lack of sufficiently intense, calibrated light sources. Such a system is used to collect spectral data on plasmas generated in electron beam diodes fielded on the RITS-6 accelerator (8-12MV, 140-200kA) at Sandia National Laboratories. On RITS, plasma light is collected through a small diameter (200 {mu}m) optical fiber and recorded on a fast streak camera at the output of a 1 meter Czerny-Turner monochromator. For this paper, a 300 W xenon short arc lamp (Oriel Model 6258) was used as the calibration source. Since the radiance of the xenon arc varies from cathode to anode, just the area around the tip of the cathode ('hotspot') was imaged onto the fiber, to produce the highest intensity output. To compensate for chromatic aberrations, the signal was optimized at each wavelength measured. Output power was measured using 10 nm bandpass interference filters and a calibrated photodetector. These measurements give power at discrete wavelengths across the spectrum, and when linearly interpolated, provide a calibration curve for the lamp. The shape of the spectrum is determined by the collective response of the optics, monochromator, and streak tube across the spectral region of interest. The ratio of the spectral curve to the measured bandpass filter curve at each wavelength produces a correction factor (Q) curve. This curve is then applied to the experimental data and the resultant spectra are given in absolute intensity units (photons/sec/cm{sup 2}/steradian/nm). Error analysis shows this method to be accurate to within +/- 20%, which represents a high level of accuracy for this type of measurement.

  18. Absolute calibration method for nanosecond-resolved, time-streaked, fiber optic light collection, spectroscopy systems

    NASA Astrophysics Data System (ADS)

    Johnston, Mark D.; Oliver, Bryan V.; Droemer, Darryl W.; Frogget, Brent; Crain, Marlon D.; Maron, Yitzhak

    2012-08-01

    This paper describes a convenient and accurate method to calibrate fast (<1 ns resolution) streaked, fiber optic light collection, spectroscopy systems. Such systems are inherently difficult to calibrate due to the lack of sufficiently intense, calibrated light sources. Such a system is used to collect spectral data on plasmas generated in electron beam diodes fielded on the RITS-6 accelerator (8-12MV, 140-200kA) at Sandia National Laboratories. On RITS, plasma light is collected through a small diameter (200 μm) optical fiber and recorded on a fast streak camera at the output of a 1 meter Czerny-Turner monochromator. For this paper, a 300 W xenon short arc lamp (Oriel Model 6258) was used as the calibration source. Since the radiance of the xenon arc varies from cathode to anode, just the area around the tip of the cathode ("hotspot") was imaged onto the fiber, to produce the highest intensity output. To compensate for chromatic aberrations, the signal was optimized at each wavelength measured. Output power was measured using 10 nm bandpass interference filters and a calibrated photodetector. These measurements give power at discrete wavelengths across the spectrum, and when linearly interpolated, provide a calibration curve for the lamp. The shape of the spectrum is determined by the collective response of the optics, monochromator, and streak tube across the spectral region of interest. The ratio of the spectral curve to the measured bandpass filter curve at each wavelength produces a correction factor (Q) curve. This curve is then applied to the experimental data and the resultant spectra are given in absolute intensity units (photons/sec/cm2/steradian/nm). Error analysis shows this method to be accurate to within +/- 20%, which represents a high level of accuracy for this type of measurement.

  19. Counting particles in tissue sections: choices of methods and importance of calibration to minimize biases.

    PubMed

    von Bartheld, C

    2002-04-01

    Investigators must choose between counting methods to quantify microscopic particles in tissues. The conventional profile-based ("model-based" or "2D-") counting methods have been criticized for their potential biases due to assumptions about shapes, sizes, and orientation of particles when converting profile counts into cell numbers. New stereological methods ("design-based" or "3D-") methods such as the optical disector or physical disector were initially introduced as being inherently unbiased. Recent calibration analyses and comparisons of results from different investigators have revealed the potential for significant biases in the most efficient and most frequently used design-based method, the optical disector. This review aims to objectively assess the strengths and limitations of current profile- and disector-based cell counting methods by examination of studies in which these methods have been calibrated against the "gold-standard", counts obtained by 3-dimensional reconstruction of serial sections. Advantages and disadvantages of each counting method and the associated embedding and sectioning techniques are compared and frequent mistakes and pitfalls of each technique are discussed. The importance of a calibration step for each technique is emphasized, and a protocol is provided for a quick and simple calibration by a "sampling" 3-D reconstruction of limited serial sections. Trends in the usage of counting methods are analyzed in four major journals. It is hoped that this review will be helpful, for both investigators and manuscript reviewers, in clarifying some of the contentious issues in the choice and implementation of appropriate methods for particle counting in tissue sections. PMID:11962763

  20. New methods for treatment effect calibration, with applications to non-inferiority trials.

    PubMed

    Zhang, Zhiwei; Nie, Lei; Soon, Guoxing; Hu, Zonghui

    2016-03-01

    In comparative effectiveness research, it is often of interest to calibrate treatment effect estimates from a clinical trial to a target population that differs from the study population. One important application is an indirect comparison of a new treatment with a placebo control on the basis of two separate randomized clinical trials: a non-inferiority trial comparing the new treatment with an active control and a historical trial comparing the active control with placebo. The available methods for treatment effect calibration include an outcome regression (OR) method based on a regression model for the outcome and a weighting method based on a propensity score (PS) model. This article proposes new methods for treatment effect calibration: one based on a conditional effect (CE) model and two doubly robust (DR) methods. The first DR method involves a PS model and an OR model, is asymptotically valid if either model is correct, and attains the semiparametric information bound if both models are correct. The second DR method involves a PS model, a CE model, and possibly an OR model, is asymptotically valid under the union of the PS and CE models, and attains the semiparametric information bound if all three models are correct. The various methods are compared in a simulation study and applied to recent clinical trials for treating human immunodeficiency virus infection. PMID:26363775

  1. Self-referencing calibration method for transmission spheres in Fizeau interferometry

    NASA Astrophysics Data System (ADS)

    Burke, Jan; Wu, David S.

    2010-08-01

    The calibration of reference surfaces becomes important in interferometry whenever the tolerances for the tested component are comparable to the imperfections of the reference surface itself. To achieve measurement accuracy better than the reference surface, its errors must be characterised and subtracted from the measurement result. We propose a rapid and simple technique utilising a flat mirror in the focus of the converging test wavefront and a partial occlusion of the test beam, to implement a double-pass self-calibration of the reference surface. Stitching together three or more measurements, with the beam stop appropriately rotated, yields the full-aperture calibration data. The method cannot detect point-antisymmetric errors, but common errors in reference spheres, such as spherical aberration and astigmatism, are point-symmetric and should still be adequately captured. For calibrating spherical surfaces in Fizeau interferometry, a ball of good sphericity can be measured against the reference surface in a number of random orientations. This averages out the errors of the ball and converges toward the stationary error in the reference sphere. Depending on the quality of the ball and the desired uncertainty, the number of orientations required can be large (50-100), which is laborious and time-consuming. We compare the performance of the new technique with the ball-averaging method and the so-called "cat's eye" method to assess the practical trade-offs involved.

  2. A velocity dependent effective angle method for calibration of X-probes at low velocities

    NASA Astrophysics Data System (ADS)

    Bakken, Ole Martin; Krogstad, Per-Åge

    A velocity dependent effective angle (VDEA) method for the calibration of yaw response of hot-wire X-probes at low flow velocities (0.5-6 m/s) is presented. Comparisons with a full velocity vs. yaw-angle method (Österlund 1999) in a smooth wall channel flow indicate that there is only moderate advantage in using the latter method, which is considerably more laborious. Comparisons with direct numerical simulations (DNS) (Moser et al. 1999) and the more common fixed effective angle method (FEA) show that the VDEA method significantly improves estimates of Reynolds stresses compared to the FEA method.

  3. Interferometric method for phase calibration in liquid crystal spatial light modulators using a self-generated diffraction-grating.

    PubMed

    Fuentes, José Luis Martínez; Fernández, Enrique J; Prieto, Pedro M; Artal, Pablo

    2016-06-27

    An auto-referenced interferometric method for calibrating phase modulation of parallel-aligned liquid crystal (PAL) spatial light modulators (SLM) is described. The method is experimentally straightforward, robust, and requires solely of a collimated beam, with no need of additional optics. This method uses the SLM itself to create a tilted plane wave and a reference wave which mutually interfere. These waves are codified by means of a binary diffraction grating and a uniformly distributed gray level area (piston) into the SLM surface. Phase shift for each gray level addressed to the piston section can then be evaluated. Phase modulation on the SLM can also be retrieved with the proposed method over spatially resolved portions of the surface. Phase information obtained with this novel method is compared to other well established calibration procedures, requiring extra elements and more elaborated optical set-ups. The results show a good agreement with previous methods. The advantages of the new method include high mechanical stability, faster performance, and a significantly easier practical implementation. PMID:27410574

  4. Validation of the onboard radiometric calibration of the GOES I-M visible channel by reflectance-based vicarious methods

    NASA Astrophysics Data System (ADS)

    Leisso, Nathan P.; Thome, Kurtis J.; Czapla-Myers, Jeffrey S.

    2007-09-01

    The current generation of the Geostationary Operations Environmental Satellite (GOES) platform employs a total of 5 sensors to monitor and record atmospheric conditions used in predictions of upcoming weather events. Included in this package is a 5-band imager that, from the 36,000-km geosynchronous orbit inhabited by GOES platform, enables multiple fixed full-disc surface images of the earth during the course of a 24-hour day. There is currently no on-board radiometric calibration for the visible bands of the imager and radiometric calibration relies on vicarious approaches. The Remote Sensing Group (RSG) at the University of Arizona uses a vicarious approach that relies on ground-based measurements to determine the radiometric calibration for multiple sun-synchronous and airborne visible and near-infrared sensors. The current work extends the approach to the GOES I-M series of sensor. The paper presents the methods and results of the reflectance-based method applied to the 1-km visible channel of GOES-11using large North American high-desert test sites. Modifications to the RSG's methods to take into account the location of the test sites at large zenith angles within the full-disk GOES image. The work provides an opportunity to evaluate uncertainties of the spectral BRF of the test sites at large view angles and resulting importance to the accurate radiometric calibration of a sensor. In addition, the impact of increased path length caused by the large view angle is evaluated with an emphasis on the increased effect of the atmospheric characterization.

  5. Methods for detecting additional genes underlying Alzheimer disease

    SciTech Connect

    Locke, P.A.; Haines, J.L.; Ter-Minassian, M.

    1994-09-01

    Alzheimer`s disease (AD) is a complex inherited disorder with proven genetic heterogeneity. To date, genes on chromosome 21 (APP) and 14 (not yet identified) are associated with early-onset familial AD, while the APOE gene on chromosome 19 is associated with both late onset familial and sporadic AD and early onset sporadic AD. Although these genes likely account for the majority of AD, many familial cases cannot be traced to any of these genes. From a set of 127 late-onset multiplex families screened for APOE, 43 (34%) families have at least one affected individual with no APOE-4 allele, suggesting an alternative genetic etiology. Simulation studies indicated that additional loci could be identified through a genomic screen with a 10 cM sieve on a subset of 21 well documented, non-APOE-4 families. Given the uncertainties in the mode of inheritance, reliance on a single analytical method could result in a missed linkage. Therefore, we have developed a strategy of using multiple overlapping yet complementary methods to detect linkage. These include sib-pair analysis and affected-pedigree-member analysis, neither of which makes assumptions about mode of inheritance, and lod score analysis (using two predefined genetic models). In order for a marker to qualify for follow-up, it must fit at least two of three criteria. These are nominal P values of 0.05 or less for the non-parametric methods, and/or a lod score greater than 1.0. Adjacent markers each fulfilling a single criterion also warrant follow-up. To date, we have screened 61 markers on chromosomes 1, 2, 3, 18, 19, 21, and 22. One marker, D2S163, generated a lod score of 1.06 ({theta} = 0.15) and an APMT statistic of 3.68 (P < 0.001). This region is currently being investigated in more detail. Updated results of this region plus additional screening data will be presented.

  6. Comparison of different methods for liquid level adjustment in tank prover calibration

    NASA Astrophysics Data System (ADS)

    Garcia, D. A.; Farias, E. C.; Gabriel, P. C.; Aquino, M. H.; Gomes, R. S. E.; Y Aibe, V.

    2015-01-01

    The adjustment of the liquid level during the calibration of tank provers with fixed volume is normally done by overfill but it can be done in different ways. In this article four level adjustment techniques are compared: plate, pipette, ruler and overfill adjustment. The adjustment methods using plate and pipette presented good agreement with the tank's nominal volume and lower uncertainty among the tested methods.

  7. A fully Bayesian method for jointly fitting instrumental calibration and X-ray spectral models

    SciTech Connect

    Xu, Jin; Yu, Yaming; Van Dyk, David A.; Kashyap, Vinay L.; Siemiginowska, Aneta; Drake, Jeremy; Ratzlaff, Pete; Connors, Alanna; Meng, Xiao-Li E-mail: yamingy@ics.uci.edu E-mail: vkashyap@cfa.harvard.edu E-mail: jdrake@cfa.harvard.edu E-mail: meng@stat.harvard.edu

    2014-10-20

    Owing to a lack of robust principled methods, systematic instrumental uncertainties have generally been ignored in astrophysical data analysis despite wide recognition of the importance of including them. Ignoring calibration uncertainty can cause bias in the estimation of source model parameters and can lead to underestimation of the variance of these estimates. We previously introduced a pragmatic Bayesian method to address this problem. The method is 'pragmatic' in that it introduced an ad hoc technique that simplified computation by neglecting the potential information in the data for narrowing the uncertainty for the calibration product. Following that work, we use a principal component analysis to efficiently represent the uncertainty of the effective area of an X-ray (or γ-ray) telescope. Here, however, we leverage this representation to enable a principled, fully Bayesian method that coherently accounts for the calibration uncertainty in high-energy spectral analysis. In this setting, the method is compared with standard analysis techniques and the pragmatic Bayesian method. The advantage of the fully Bayesian method is that it allows the data to provide information not only for estimation of the source parameters but also for the calibration product—here the effective area, conditional on the adopted spectral model. In this way, it can yield more accurate and efficient estimates of the source parameters along with valid estimates of their uncertainty. Provided that the source spectrum can be accurately described by a parameterized model, this method allows rigorous inference about the effective area by quantifying which possible curves are most consistent with the data.

  8. Development and evaluation of a method of calibrating medical displays based on fixed adaptation

    SciTech Connect

    Sund, Patrik Månsson, Lars Gunnar; Båth, Magnus

    2015-04-15

    Purpose: The purpose of this work was to develop and evaluate a new method for calibration of medical displays that includes the effect of fixed adaptation and by using equipment and luminance levels typical for a modern radiology department. Methods: Low contrast sinusoidal test patterns were derived at nine luminance levels from 2 to 600 cd/m{sup 2} and used in a two alternative forced choice observer study, where the adaptation level was fixed at the logarithmic average of 35 cd/m{sup 2}. The contrast sensitivity at each luminance level was derived by establishing a linear relationship between the ten pattern contrast levels used at every luminance level and a detectability index (d′) calculated from the fraction of correct responses. A Gaussian function was fitted to the data and normalized to the adaptation level. The corresponding equation was used in a display calibration method that included the grayscale standard display function (GSDF) but compensated for fixed adaptation. In the evaluation study, the contrast of circular objects with a fixed pixel contrast was displayed using both calibration methods and was rated on a five-grade scale. Results were calculated using a visual grading characteristics method. Error estimations in both observer studies were derived using a bootstrap method. Results: The contrast sensitivities for the darkest and brightest patterns compared to the contrast sensitivity at the adaptation luminance were 37% and 56%, respectively. The obtained Gaussian fit corresponded well with similar studies. The evaluation study showed a higher degree of equally distributed contrast throughout the luminance range with the calibration method compensated for fixed adaptation than for the GSDF. The two lowest scores for the GSDF were obtained for the darkest and brightest patterns. These scores were significantly lower than the lowest score obtained for the compensated GSDF. For the GSDF, the scores for all luminance levels were statistically

  9. A novel method to calibrate DOI function of a PET detector with a dual-ended-scintillator readout.

    PubMed

    Shao, Yiping; Yao, Rutao; Ma, Tianyu

    2008-12-01

    addition, the new method can be generally applied to calibrating other types of detectors that use the similar dual-ended readout to acquire the radiation interaction position. PMID:19175139

  10. A novel method to calibrate DOI function of a PET detector with a dual-ended-scintillator readout

    SciTech Connect

    Shao Yiping; Yao Rutao; Ma Tianyu

    2008-12-15

    addition, the new method can be generally applied to calibrating other types of detectors that use the similar dual-ended readout to acquire the radiation interaction position.

  11. Non-linearity measurements of solar cells with an LED-based combinatorial flux addition method

    NASA Astrophysics Data System (ADS)

    Hamadani, Behrang H.; Shore, Andrew; Roller, John; Yoon, Howard W.; Campanelli, Mark

    2016-02-01

    We present a light emitting diode (LED)-based system utilizing a combinatorial flux addition method to investigate the non-linear relationship in solar cells between the output current of the cell and the incident irradiance level. The magnitude of the light flux is controlled by the supplied currents to two LEDs (or two sets of them) in a combinatorial fashion. The signals measured from the cell are arranged within a related overdetermined linear system of equations derived from an appropriately chosen Nth degree polynomial representing the relationship between the measured signals and the incident fluxes. The flux values and the polynomial coefficients are then solved for by linear least squares to obtain the best fit. The technique can be applied to any solar cell, under either monochromatic or broadband spectrum. For the unscaled solution, no reference detectors or prior calibrations of the light flux are required. However, if at least one calibrated irradiance value is known, then the entire curve can be scaled to an appropriate spectral responsivity value. Using this technique, a large number of data points can be obtained in a relatively short time scale over a large signal range.

  12. A calibration method for proposed XRF measurements of arsenic and selenium in nail clippings.

    PubMed

    Gherase, Mihai R; Fleming, David E B

    2011-10-21

    A calibration method for proposed x-ray fluorescence (XRF) measurements of arsenic and selenium in nail clippings is demonstrated. Phantom nail clippings were produced from a whole nail phantom (0.7 mm thickness, 25 × 25 mm(2) area) and contained equal concentrations of arsenic and selenium ranging from 0 to 20 µg g(-1) in increments of 5 µg g(-1). The phantom nail clippings were then grouped in samples of five different masses: 20, 40, 60, 80 and 100 mg for each concentration. Experimental x-ray spectra were acquired for each of the sample masses using a portable x-ray tube and a detector unit. Calibration lines (XRF signal in a number of counts versus stoichiometric elemental concentration) were produced for each of the two elements. A semi-empirical relationship between the mass of the nail phantoms (m) and the slope of the calibration line (s) was determined separately for arsenic and selenium. Using this calibration method, one can estimate elemental concentrations and their uncertainties from the XRF spectra of human nail clippings. PMID:21937772

  13. Analytic method based on identification of ellipse parameters for scanner calibration in cone-beam tomography.

    PubMed

    Noo, F; Clackdoyle, R; Mennessier, C; White, T A; Roney, T J

    2000-11-01

    This paper is about calibration of cone-beam (CB) scanners for both x-ray computed tomography and single-photon emission computed tomography. Scanner calibration refers here to the estimation of a set of parameters which fully describe the geometry of data acquisition. Such parameters are needed for the tomographic reconstruction step. The discussion is limited to the usual case where the cone vertex and planar detector move along a circular path relative to the object. It is also assumed that the detector does not have spatial distortions. We propose a new method which requires a small set of measurements of a simple calibration object consisting of two spherical objects, that can be considered as 'point' objects. This object traces two ellipses on the detector and from the parametric description of these ellipses, the calibration geometry can be determined analytically using explicit formulae. The method is robust and easy to implement. However, it is not fully general as it is assumed that the detector is parallel to the rotation axis of the scanner. Implementation details are given for an experimental x-ray CB scanner. PMID:11098919

  14. A calibration method for proposed XRF measurements of arsenic and selenium in nail clippings

    NASA Astrophysics Data System (ADS)

    Gherase, Mihai R.; Fleming, David E. B.

    2011-10-01

    A calibration method for proposed x-ray fluorescence (XRF) measurements of arsenic and selenium in nail clippings is demonstrated. Phantom nail clippings were produced from a whole nail phantom (0.7 mm thickness, 25 × 25 mm2 area) and contained equal concentrations of arsenic and selenium ranging from 0 to 20 µg g-1 in increments of 5 µg g-1. The phantom nail clippings were then grouped in samples of five different masses: 20, 40, 60, 80 and 100 mg for each concentration. Experimental x-ray spectra were acquired for each of the sample masses using a portable x-ray tube and a detector unit. Calibration lines (XRF signal in a number of counts versus stoichiometric elemental concentration) were produced for each of the two elements. A semi-empirical relationship between the mass of the nail phantoms (m) and the slope of the calibration line (s) was determined separately for arsenic and selenium. Using this calibration method, one can estimate elemental concentrations and their uncertainties from the XRF spectra of human nail clippings.

  15. [Research on absolute calibration of sun channel of sun photometer using laser raster scanning method].

    PubMed

    Xu, Wen-Bin; Li, Jian-Jun; Zheng, Xiao-Bing

    2013-01-01

    In the present paper, a new calibration method of absolute spectral irradiance responsivity of sun channel of sun photometer was developed. A tunable laser was used as source and a standard tranfer detector, calibrated against cryogenic absolute radiometer, was used to measure laser beam power. By raster scanning of a single collimated laser beam to generate the uniform irradiance field at the plane of effective aperture stop of sun photometer, the absolute irradiance responsivity of center wavelength of the 870 nm unpolarized sun channels of sun photometer was obtained accurately. The relative spectral irradiance responsivity of corresponding channel was obtained by using lamp-monochromator system and then used to acquire the absolute spectral irradiance responsivity in the laboratory. On the basis of the above results, the top-of-the-atmosphere responsive constant V0 was obtained by integration with extraterrestrial solar spectral irradiance data. Comparing the calibration result with that from GSFC, NASA in 2009, the difference is only 3.75%. In the last, the uncertainties of calibration were evaluated and reached to 2.06%. The principle feasibility of the new method was validated. PMID:23586268

  16. Hybrid PSO-ASVR-based method for data fitting in the calibration of infrared radiometer

    NASA Astrophysics Data System (ADS)

    Yang, Sen; Li, Chengwei

    2016-06-01

    The present paper describes a hybrid particle swarm optimization-adaptive support vector regression (PSO-ASVR)-based method for data fitting in the calibration of infrared radiometer. The proposed hybrid PSO-ASVR-based method is based on PSO in combination with Adaptive Processing and Support Vector Regression (SVR). The optimization technique involves setting parameters in the ASVR fitting procedure, which significantly improves the fitting accuracy. However, its use in the calibration of infrared radiometer has not yet been widely explored. Bearing this in mind, the PSO-ASVR-based method, which is based on the statistical learning theory, is successfully used here to get the relationship between the radiation of a standard source and the response of an infrared radiometer. Main advantages of this method are the flexible adjustment mechanism in data processing and the optimization mechanism in a kernel parameter setting of SVR. Numerical examples and applications to the calibration of infrared radiometer are performed to verify the performance of PSO-ASVR-based method compared to conventional data fitting methods.

  17. Hybrid PSO-ASVR-based method for data fitting in the calibration of infrared radiometer.

    PubMed

    Yang, Sen; Li, Chengwei

    2016-06-01

    The present paper describes a hybrid particle swarm optimization-adaptive support vector regression (PSO-ASVR)-based method for data fitting in the calibration of infrared radiometer. The proposed hybrid PSO-ASVR-based method is based on PSO in combination with Adaptive Processing and Support Vector Regression (SVR). The optimization technique involves setting parameters in the ASVR fitting procedure, which significantly improves the fitting accuracy. However, its use in the calibration of infrared radiometer has not yet been widely explored. Bearing this in mind, the PSO-ASVR-based method, which is based on the statistical learning theory, is successfully used here to get the relationship between the radiation of a standard source and the response of an infrared radiometer. Main advantages of this method are the flexible adjustment mechanism in data processing and the optimization mechanism in a kernel parameter setting of SVR. Numerical examples and applications to the calibration of infrared radiometer are performed to verify the performance of PSO-ASVR-based method compared to conventional data fitting methods. PMID:27370427

  18. A method to calibrate phase fluctuation in polarization-sensitive swept-source optical coherence tomography

    NASA Astrophysics Data System (ADS)

    Lu, Zenghai; Kasaragod, Deepa K.; Matcher, Stephen J.

    2011-06-01

    A phase fluctuation calibration method is presented for polarization-sensitive swept-source optical coherence tomography (PS-SS-OCT) using continuous polarization modulation. The method consists of the generation of a continuous triggered tone-burst waveform rather than an asynchronous waveform by use of a function generator and the removal of the global phases of the measured Jones matrices by use of matrix normalization. This could remove the use of auxiliary optical components for the phase fluctuation compensation in the system, which reduces the system complexity. Phase fluctuation calibration is necessary to obtain the reference Jones matrix by averaging the measured Jones matrices at sample surfaces. Measurements on an equine tendon sample were made by the PS-SS-OCT system to validate the proposed method.

  19. A Focusing Method in the Calibration Process of Image Sensors Based on IOFBs

    PubMed Central

    Fernández, Pedro R.; Lázaro, José L.; Gardel, Alfredo; Cano, Ángel E.; Bravo, Ignacio

    2010-01-01

    A focusing procedure in the calibration process of image sensors based on Incoherent Optical Fiber Bundles (IOFBs) is described using the information extracted from fibers. These procedures differ from any other currently known focusing method due to the non spatial in-out correspondence between fibers, which produces a natural codification of the image to transmit. Focus measuring is essential prior to carrying out calibration in order to guarantee accurate processing and decoding. Four algorithms have been developed to estimate the focus measure; two methods based on mean grey level, and the other two based on variance. In this paper, a few simple focus measures are defined and compared. Some experimental results referred to the focus measure and the accuracy of the developed methods are discussed in order to demonstrate its effectiveness. PMID:22315526

  20. An improved method of energy calibration for position-sensitive silicon detectors

    NASA Astrophysics Data System (ADS)

    Sun, Ming-Dao; Huang, Tian-Heng; Liu, Zhong; Ding, Bing; Yang, Hua-Bin; Zhang, Zhi-Yuan; Wang, Jian-Guo; Ma, Long; Yu, Lin; Wang, Yong-Sheng; Gan, Zai-Guo; Xiao-Hong, Zhou

    2016-04-01

    Energy calibration of resistive charge division-based position-sensitive silicon detectors is achieved by parabolic fitting in the traditional method, where the systematic variations of vertex and curvature of the parabola with energy must be considered. In this paper we extend the traditional method in order to correct the fitting function, simplify the procedure of calibration and improve the experimental data quality. Instead of a parabolic function as used in the traditional method, a new function describing the relation of position and energy is introduced. The energy resolution of the 8.088 MeV α decay of 213Rn is determined to be about 87 keV (FWHM), which is better than the result of the traditional method, 104 keV (FWHM). The improved method can be applied to the energy calibration of resistive charge division-based position-sensitive silicon detectors with various performances. Supported by ‘100 Person Project’ of the Chinese Academy of Sciences and the National Natural Science Foundation of China (11405224 and 11435014)

  1. A Global Calibration Method for Widely Distributed Cameras Based on Vanishing Features

    PubMed Central

    Wu, Xiaolong; Wu, Sentang; Xing, Zhihui; Jia, Xiang

    2016-01-01

    This paper presents a global calibration method for widely distributed vision sensors in ring-topologies. Planar target with two mutually orthogonal groups of parallel lines is needed for each camera. Firstly, the relative pose of each camera and its corresponding target is found from the vanishing points and lines. Next, an auxiliary camera is used to find the relative poses between neighboring pairs of calibration targets. Then the relative pose from each target to the reference target is initialized by the chain of transformations, followed by nonlinear optimization based on the constraint of ring-topologies. Lastly, the relative poses between the cameras are found from the relative poses of calibration targets. Synthetic data, simulation images and real experiments all demonstrate that the proposed method is reliable and accurate. The accumulated error due to multiple coordinate transformations can be adjusted effectively by the proposed method. In real experiment, eight targets are located in an area about 1200 mm × 1200 mm. The accuracy of the proposed method is about 0.465 mm when the times of coordinate transformations reach a maximum. The proposed method is simple and can be applied to different camera configurations. PMID:27338386

  2. A shearing-based method for the simultaneous calibration of angle measuring devices

    NASA Astrophysics Data System (ADS)

    Geckeler, Ralf D.; Just, Andreas

    2014-10-01

    In this paper, we present a novel adaptation of the shearing method to the simultaneous calibration of angle measuring devices which is based on multiple comparisons of their angle readings in different relative angular orientations. Without recourse to an external standard, the errors of the two devices can be recovered, up to their linear components, from a set of three comparisons. We demonstrate the method by the use of an autocollimator and an angle encoder. It proved to be ideally suited for the calibration of interpolation errors of the devices at small angular scales which are difficult to characterize with other methods. In the case that the linear components of the errors are needed, too, only two angle differences, which correspond to the changes in the relative angular orientations of the devices, need to be traced back to an external standard. A comprehensive overview, both theoretical and experimental, of the capabilities and limitations of the method is presented, including experimental data obtained with the high-precision primary angle standard of PTB. We demonstrate error-separation with a standard measurement uncertainty at a level of 1 milliarcsecond (5 nrad) which, when compared to uncertainties reachable by conventional calibration methods for autocollimators, represents an improvement by a factor of 2-3.

  3. Method for controlling a laser additive process using intrinsic illumination

    NASA Astrophysics Data System (ADS)

    Tait, Robert; Cai, Guoshuang; Azer, Magdi; Chen, Xiaobin; Liu, Yong; Harding, Kevin

    2015-05-01

    One form of additive manufacturing is to use a laser to generate a melt pool from powdered metal that is sprayed from a nozzle. The laser net-shape machining system builds the part a layer at a time by following a predetermined path. However, because the path may need to take many turns, maintaining a constant melt pool may not be easy. A straight section may require one speed and power while a sharp bend would over melt the metal at the same settings. This paper describes a process monitoring method that uses the intrinsic IR radiation from the melt pool along with a process model configured to establish target values for the parameters associated with the manufacture or repair. This model is based upon known properties of the metal being used as well as the properties of the laser beam. An adaptive control technique is then employed to control process parameters of the machining system based upon the real-time weld pool measurement. Since the system uses the heat radiant from the melt pool, other previously deposited metal does not confuse the system as only the melted material is seen by the camera.

  4. Influence of calibration method and material on the accuracy of stress distribution measurement systems.

    PubMed

    Engel, Karsten; Hartmann, Ulrich; Potthast, Wolfgang; Brüggemann, Gert-Peter

    2016-06-01

    Biomechanical analyses of the stress distribution and the force transfer in the human knee are essential to better understand the aetiology of joint diseases. Accuracy studies of commonly used capacitive or resistive-based stress distribution measurement systems have led to severe problems caused by an inaccurate experimental setup. For instance, in one study, overestimations of the measured forces in the sensor's centre were reported. Therefore, the primary aim of this study was to investigate the ability of capacitive and resistive-based sensors to measure forces in a homogenous pressure environment and the secondary goal was to analyse the influence of different calibration materials on the measurement accuracy. A Novel pressure vessel and metal indenters covered with different rubber materials were used in combination with a material testing machine to load the sensors. Four different linearly increasing nominal forces (925-3670 N) were applied and the deviations between the nominal and the measured forces were calculated. The capacitive measurement system showed errors between 1% and 7% in the homogenous pressure environment, whereas the errors of the resistive system were found to vary between 4% and 17%. The influence of the calibration material was observed to be greater for the resistive sensors (1-179%) than for the capacitive sensors (0.5-25%). In conclusion, it can be stated that - for the pressure measurement systems compared in this article - the capacitive one is less sensitive to the calibration method and the calibration material than the resistive system. PMID:26146092

  5. Optical Calibration Process Developed for Neural-Network-Based Optical Nondestructive Evaluation Method

    NASA Technical Reports Server (NTRS)

    Decker, Arthur J.

    2004-01-01

    A completely optical calibration process has been developed at Glenn for calibrating a neural-network-based nondestructive evaluation (NDE) method. The NDE method itself detects very small changes in the characteristic patterns or vibration mode shapes of vibrating structures as discussed in many references. The mode shapes or characteristic patterns are recorded using television or electronic holography and change when a structure experiences, for example, cracking, debonds, or variations in fastener properties. An artificial neural network can be trained to be very sensitive to changes in the mode shapes, but quantifying or calibrating that sensitivity in a consistent, meaningful, and deliverable manner has been challenging. The standard calibration approach has been difficult to implement, where the response to damage of the trained neural network is compared with the responses of vibration-measurement sensors. In particular, the vibration-measurement sensors are intrusive, insufficiently sensitive, and not numerous enough. In response to these difficulties, a completely optical alternative to the standard calibration approach was proposed and tested successfully. Specifically, the vibration mode to be monitored for structural damage was intentionally contaminated with known amounts of another mode, and the response of the trained neural network was measured as a function of the peak-to-peak amplitude of the contaminating mode. The neural network calibration technique essentially uses the vibration mode shapes of the undamaged structure as standards against which the changed mode shapes are compared. The published response of the network can be made nearly independent of the contaminating mode, if enough vibration modes are used to train the net. The sensitivity of the neural network can be adjusted for the environment in which the test is to be conducted. The response of a neural network trained with measured vibration patterns for use on a vibration isolation

  6. Ozone Correction for AM0 Calibrated Solar Cells for the Aircraft Method

    NASA Technical Reports Server (NTRS)

    Snyder, David B.; Scheiman, David A.; Jenkins, Phillip P.; Lyons, Valerie J. (Technical Monitor)

    2002-01-01

    The aircraft solar cell calibration method has provided cells calibrated to space conditions for 37 years. However, it is susceptible to systematic errors due to ozone concentration in the stratosphere. The present correction procedure applies a 1% increase to the measured Isc values. High band-gap cells are more sensitive to ozone adsorbed wavelengths so it has become important to reassess the correction technique. This paper evaluates the ozone correction to be 1+{O3}sup Fo, where Fo is 29.5x10(exp-6)/d.u. for a Silicon solar cell and 42.2xl0(exp -6)/d.u. for a GaAs cell. Results will be presented for high band-gap cells. A comparison with flight data indicates that this method of correcting for the ozone density improves the uncertainty of AM0 Isc to 0.5%.

  7. A High Speed Calibration Method for Laser Positioner by Constant Velocity Scanning

    NASA Astrophysics Data System (ADS)

    Ono, Hiroyuki

    This paper describes a high speed calibration method for laser positioner by scanning work area under constant velocity. Laser positioner consists of sinusoidal laser encoder, DC motor and the controller. The encoder has diffraction grating scale and laser optics. It generates sine and cosine outputs according to the travel distance. Qualities of the scale and the optics contribute to the fluctuation of output signal amplitude, DC offset and relative phase. First, the controller collects the output data under constant velocity of 4 samples per 1 sine wave length. The parameter can be estimated by simplified discrete Fourier transform method. Calibration data are collected every quarter sine wave length over all positioner work area. The result is stored in the table then referred by the controller in real time operation. Experiment results are also reported for HDD servo track writer application.

  8. Effects of sensor calibration, balancing and parametrization on the signal space separation method

    NASA Astrophysics Data System (ADS)

    Nurminen, J.; Taulu, S.; Okada, Y.

    2008-04-01

    Signal space separation (SSS) is a novel method for processing multichannel biomagnetic data. It is useful for a variety of applications including interference suppression, movement compensation and conversion of measurements between sensor arrays. The performance of SSS has been examined mainly on a 306-channel whole-head magnetoencephalography system. To facilitate the adaptation of the method to other biomagnetometer systems, the effect of various properties of the sensor array on its performance needs to be studied. To this end, we examined the effects of gradiometer imbalance, sensor calibration errors and erroneous sensor geometry information on SSS using simulations. The results indicate that depending on the application, gradiometer balance on the level of 0.1% to 0.5% may be needed for satisfactory SSS performance. For wire-wound gradiometers, this requires very careful attention in manufacturing. Errors in calibration coefficients and geometry information were found to have less significance.

  9. Effects of sensor calibration, balancing and parametrization on the signal space separation method.

    PubMed

    Nurminen, J; Taulu, S; Okada, Y

    2008-04-01

    Signal space separation (SSS) is a novel method for processing multichannel biomagnetic data. It is useful for a variety of applications including interference suppression, movement compensation and conversion of measurements between sensor arrays. The performance of SSS has been examined mainly on a 306-channel whole-head magnetoencephalography system. To facilitate the adaptation of the method to other biomagnetometer systems, the effect of various properties of the sensor array on its performance needs to be studied. To this end, we examined the effects of gradiometer imbalance, sensor calibration errors and erroneous sensor geometry information on SSS using simulations. The results indicate that depending on the application, gradiometer balance on the level of 0.1% to 0.5% may be needed for satisfactory SSS performance. For wire-wound gradiometers, this requires very careful attention in manufacturing. Errors in calibration coefficients and geometry information were found to have less significance. PMID:18354243

  10. Method for in-situ restoration of platinum resistance thermometer calibration

    SciTech Connect

    Carroll, R.M.

    1989-01-03

    A method is described for the in situ restoration of a platinum resistance thermometer, which has become decalibrated due to oxide surface film formation and/or strain-related damage, to a stress-free calibration condition wherein the thermometer includes a platinum resistor sensing element whose resistance varies with the temperature of the element in accordance with a known relationship. The method consists of: passing a controlled dc current through the platinum resistor sensing element, the current having a magnitude sufficient to raise the temperature of the element to its oxide decomposition and annealing temperature. The current is maintained for a heating period sufficient to restore the element to an oxide free surface and stress-free calibration condition.

  11. Calibration method and apparatus for measuring the concentration of components in a fluid

    DOEpatents

    Durham, Michael D.; Sagan, Francis J.; Burkhardt, Mark R.

    1993-01-01

    A calibration method and apparatus for use in measuring the concentrations of components of a fluid is provided. The measurements are determined from the intensity of radiation over a selected range of radiation wavelengths using peak-to-trough calculations. The peak-to-trough calculations are simplified by compensating for radiation absorption by the apparatus. The invention also allows absorption characteristics of an interfering fluid component to be accurately determined and negated thereby facilitating analysis of the fluid.

  12. Calibration method and apparatus for measuring the concentration of components in a fluid

    DOEpatents

    Durham, M.D.; Sagan, F.J.; Burkhardt, M.R.

    1993-12-21

    A calibration method and apparatus for use in measuring the concentrations of components of a fluid is provided. The measurements are determined from the intensity of radiation over a selected range of radiation wavelengths using peak-to-trough calculations. The peak-to-trough calculations are simplified by compensating for radiation absorption by the apparatus. The invention also allows absorption characteristics of an interfering fluid component to be accurately determined and negated thereby facilitating analysis of the fluid. 7 figures.

  13. Application of Response Surface based Calibration and Sensitivity Analysis methods for Regional Hydrogeological Modelling in the Western Canada Sedimentary Basin

    NASA Astrophysics Data System (ADS)

    Singh, A.; Palombi, D.; Huff, G. F.

    2014-12-01

    A regional scale study of groundwater flow dynamics was undertaken in the Western Canada Sedimentary Basin (WCSB), comprising parts of Alberta, Saskatchewan and British Columbia. The objective of the study is to investigate basin-scale hydrogeology in WCSB and to establish boundary conditions for future local-scale groundwater management models. Earlier work in the Alberta basin has acknowledged the fact that in addition to topography controlled conditions, a substantial part of the basin exhibits sub-hydrostatic regimes. The basin-scale model (approx. 420,000 km2) includes Upper Cretaceous aquifers to Recent age sediments which collectively attain maximum thicknesses of >2600 m. Regional aquifer units considered for the numerical model are Quaternary sediments, and the sedimentary rocks of the Paskapoo, Scollard, Horseshoe Canyon formations and the Belly River Group. Regional aquitards delineated include the Battle and Bear Paw formations. The study area is bound to the west by the Brazeau-Waptiti thrust belt and to the south by the Canada-USA international border. The boundary to the north and east is delineated by the maximum extent of the Wapiti and Belly River groups and Judith River Formation. USGS MODFLOW was implemented for numerical simulation. The steady state numerical model was calibrated using a Response Surface based (Radial Basis Functions) optimization method. The calibration targets (~2000) were comprised of drill stem tests for deeper units and static water levels for shallower units. Petrophysical analyses of cores averaged K values from analyses of aquifer test results,and literature values were used to provide initial values and calibration ranges for hydraulic properties. Results indicate predominance of topography driven, local- to intermediate-scale flow systems in all hydrostratigraphic units with recharge of these units occurring in the foothills of the Rocky Mountains. The Battle aquitard, where present, acts to retard regional flow

  14. Regression calibration method for correcting measurement-error bias in nutritional epidemiology.

    PubMed

    Spiegelman, D; McDermott, A; Rosner, B

    1997-04-01

    Regression calibration is a statistical method for adjusting point and interval estimates of effect obtained from regression models commonly used in epidemiology for bias due to measurement error in assessing nutrients or other variables. Previous work developed regression calibration for use in estimating odds ratios from logistic regression. We extend this here to estimating incidence rate ratios from Cox proportional hazards models and regression slopes from linear-regression models. Regression calibration is appropriate when a gold standard is available in a validation study and a linear measurement error with constant variance applies or when replicate measurements are available in a reliability study and linear random within-person error can be assumed. In this paper, the method is illustrated by correction of rate ratios describing the relations between the incidence of breast cancer and dietary intakes of vitamin A, alcohol, and total energy in the Nurses' Health Study. An example using linear regression is based on estimation of the relation between ultradistal radius bone density and dietary intakes of caffeine, calcium, and total energy in the Massachusetts Women's Health Study. Software implementing these methods uses SAS macros. PMID:9094918

  15. A vision-based self-calibration method for robotic visual inspection systems.

    PubMed

    Yin, Shibin; Ren, Yongjie; Zhu, Jigui; Yang, Shourui; Ye, Shenghua

    2013-01-01

    A vision-based robot self-calibration method is proposed in this paper to evaluate the kinematic parameter errors of a robot using a visual sensor mounted on its end-effector. This approach could be performed in the industrial field without external, expensive apparatus or an elaborate setup. A robot Tool Center Point (TCP) is defined in the structural model of a line-structured laser sensor, and aligned to a reference point fixed in the robot workspace. A mathematical model is established to formulate the misalignment errors with kinematic parameter errors and TCP position errors. Based on the fixed point constraints, the kinematic parameter errors and TCP position errors are identified with an iterative algorithm. Compared to the conventional methods, this proposed method eliminates the need for a robot-based-frame and hand-to-eye calibrations, shortens the error propagation chain, and makes the calibration process more accurate and convenient. A validation experiment is performed on an ABB IRB2400 robot. An optimal configuration on the number and distribution of fixed points in the robot workspace is obtained based on the experimental results. Comparative experiments reveal that there is a significant improvement of the measuring accuracy of the robotic visual inspection system. PMID:24300597

  16. A Vision-Based Self-Calibration Method for Robotic Visual Inspection Systems

    PubMed Central

    Yin, Shibin; Ren, Yongjie; Zhu, Jigui; Yang, Shourui; Ye, Shenghua

    2013-01-01

    A vision-based robot self-calibration method is proposed in this paper to evaluate the kinematic parameter errors of a robot using a visual sensor mounted on its end-effector. This approach could be performed in the industrial field without external, expensive apparatus or an elaborate setup. A robot Tool Center Point (TCP) is defined in the structural model of a line-structured laser sensor, and aligned to a reference point fixed in the robot workspace. A mathematical model is established to formulate the misalignment errors with kinematic parameter errors and TCP position errors. Based on the fixed point constraints, the kinematic parameter errors and TCP position errors are identified with an iterative algorithm. Compared to the conventional methods, this proposed method eliminates the need for a robot-based-frame and hand-to-eye calibrations, shortens the error propagation chain, and makes the calibration process more accurate and convenient. A validation experiment is performed on an ABB IRB2400 robot. An optimal configuration on the number and distribution of fixed points in the robot workspace is obtained based on the experimental results. Comparative experiments reveal that there is a significant improvement of the measuring accuracy of the robotic visual inspection system. PMID:24300597

  17. THE EURADOS-KIT TRAINING COURSE ON MONTE CARLO METHODS FOR THE CALIBRATION OF BODY COUNTERS.

    PubMed

    Breustedt, B; Broggio, D; Gomez-Ros, J M; Leone, D; Marzocchi, O; Poelz, S; Shutt, A; Lopez, M A

    2016-09-01

    Monte Carlo (MC) methods are numerical simulation techniques that can be used to extend the scope of calibrations performed in in vivo monitoring laboratories. These methods allow calibrations to be carried out for a much wider range of body shapes and sizes than would be feasible using physical phantoms. Unfortunately, nowadays, this powerful technique is still used mainly in research institutions only. In 2013, EURADOS and the in vivo monitoring laboratory of Karlsruhe Institute of Technology (KIT) organized a 3-d training course to disseminate knowledge on the application of MC methods for in vivo monitoring. It was intended as a hands-on course centered around an exercise which guided the participants step by step through the calibration process using a simplified version of KIT's equipment. Only introductory lectures on in vivo monitoring and voxel models were given. The course was based on MC codes of the MCNP family, widespread in the community. The strong involvement of the participants and the working atmosphere in the classroom as well as the formal evaluation of the course showed that the approach chosen was appropriate. Participants liked the hands-on approach and the extensive course materials on the exercise. PMID:27103642

  18. Flight Test Results of a GPS-Based Pitot-Static Calibration Method Using Output-Error Optimization for a Light Twin-Engine Airplane

    NASA Technical Reports Server (NTRS)

    Martos, Borja; Kiszely, Paul; Foster, John V.

    2011-01-01

    As part of the NASA Aviation Safety Program (AvSP), a novel pitot-static calibration method was developed to allow rapid in-flight calibration for subscale aircraft while flying within confined test areas. This approach uses Global Positioning System (GPS) technology coupled with modern system identification methods that rapidly computes optimal pressure error models over a range of airspeed with defined confidence bounds. This method has been demonstrated in subscale flight tests and has shown small 2- error bounds with significant reduction in test time compared to other methods. The current research was motivated by the desire to further evaluate and develop this method for full-scale aircraft. A goal of this research was to develop an accurate calibration method that enables reductions in test equipment and flight time, thus reducing costs. The approach involved analysis of data acquisition requirements, development of efficient flight patterns, and analysis of pressure error models based on system identification methods. Flight tests were conducted at The University of Tennessee Space Institute (UTSI) utilizing an instrumented Piper Navajo research aircraft. In addition, the UTSI engineering flight simulator was used to investigate test maneuver requirements and handling qualities issues associated with this technique. This paper provides a summary of piloted simulation and flight test results that illustrates the performance and capabilities of the NASA calibration method. Discussion of maneuver requirements and data analysis methods is included as well as recommendations for piloting technique.

  19. A calibration-independent method for accurate complex permittivity determination of liquid materials

    SciTech Connect

    Hasar, U. C.

    2008-08-15

    This note presents a calibration-independent method for accurate complex permittivity determination of liquid materials. There are two main advantages of the proposed method over those in the literature, which require measurements of two cells with different lengths loaded by the same liquid material. First, it eliminates any inhomogeneity or impurity present in the second sample and decreases the uncertainty in sample thickness. Second, it removes the undesired impacts of measurement plane deterioration on measurements of liquid materials. For validation of the proposed method, we measure the complex permittivity of distilled water and compare its extracted permittivity with the theoretical datum obtained from the Debye equation.

  20. Accurate and automatic extrinsic calibration method for blade measurement system integrated by different optical sensors

    NASA Astrophysics Data System (ADS)

    He, Wantao; Li, Zhongwei; Zhong, Kai; Shi, Yusheng; Zhao, Can; Cheng, Xu

    2014-11-01

    Fast and precise 3D inspection system is in great demand in modern manufacturing processes. At present, the available sensors have their own pros and cons, and hardly exist an omnipotent sensor to handle the complex inspection task in an accurate and effective way. The prevailing solution is integrating multiple sensors and taking advantages of their strengths. For obtaining a holistic 3D profile, the data from different sensors should be registrated into a coherent coordinate system. However, some complex shape objects own thin wall feather such as blades, the ICP registration method would become unstable. Therefore, it is very important to calibrate the extrinsic parameters of each sensor in the integrated measurement system. This paper proposed an accurate and automatic extrinsic parameter calibration method for blade measurement system integrated by different optical sensors. In this system, fringe projection sensor (FPS) and conoscopic holography sensor (CHS) is integrated into a multi-axis motion platform, and the sensors can be optimally move to any desired position at the object's surface. In order to simple the calibration process, a special calibration artifact is designed according to the characteristics of the two sensors. An automatic registration procedure based on correlation and segmentation is used to realize the artifact datasets obtaining by FPS and CHS rough alignment without any manual operation and data pro-processing, and then the Generalized Gauss-Markoff model is used to estimate the optimization transformation parameters. The experiments show the measurement result of a blade, where several sampled patches are merged into one point cloud, and it verifies the performance of the proposed method.

  1. Semi-automated calibration method for modelling of mountain permafrost evolution in Switzerland

    NASA Astrophysics Data System (ADS)

    Marmy, A.; Rajczak, J.; Delaloye, R.; Hilbich, C.; Hoelzle, M.; Kotlarski, S.; Lambiel, C.; Noetzli, J.; Phillips, M.; Salzmann, N.; Staub, B.; Hauck, C.

    2015-09-01

    Permafrost is a widespread phenomenon in the European Alps. Many important topics such as the future evolution of permafrost related to climate change and the detection of permafrost related to potential natural hazards sites are of major concern to our society. Numerical permafrost models are the only tools which facilitate the projection of the future evolution of permafrost. Due to the complexity of the processes involved and the heterogeneity of Alpine terrain, models must be carefully calibrated and results should be compared with observations at the site (borehole) scale. However, a large number of local point data are necessary to obtain a broad overview of the thermal evolution of mountain permafrost over a larger area, such as the Swiss Alps, and the site-specific model calibration of each point would be time-consuming. To face this issue, this paper presents a semi-automated calibration method using the Generalized Likelihood Uncertainty Estimation (GLUE) as implemented in a 1-D soil model (CoupModel) and applies it to six permafrost sites in the Swiss Alps prior to long-term permafrost evolution simulations. We show that this automated calibration method is able to accurately reproduce the main thermal condition characteristics with some limitations at sites with unique conditions such as 3-D air or water circulation, which have to be calibrated manually. The calibration obtained was used for RCM-based long-term simulations under the A1B climate scenario specifically downscaled at each borehole site. The projection shows general permafrost degradation with thawing at 10 m, even partially reaching 20 m depths until the end of the century, but with different timing among the sites. The degradation is more rapid at bedrock sites whereas ice-rich sites with a blocky surface cover showed a reduced sensitivity to climate change. The snow cover duration is expected to be reduced drastically (between -20 to -37 %) impacting the ground thermal regime. However

  2. Multivariate versus classical univariate calibration methods for spectrofluorimetric data: application to simultaneous determination of olmesartan medoxamil and amlodipine besylate in their combined dosage form.

    PubMed

    Darwish, Hany W; Backeit, Ahmed H

    2013-01-01

    Olmesartan medoxamil (OLM, an angiotensin II receptor blocker) and amlodipine besylate (AML, a dihydropyridine calcium channel blocker), are co-formulated in a single-dose combination for the treatment of hypertensive patients whose blood pressure is not adequately controlled on either component monotherapy. In this work, four multivariate and two univariate calibration methods were applied for simultaneous spectrofluorimetric determination of OLM and AML in their combined pharmaceutical tablets in all ratios approved by FDA. The four multivariate methods are partial least squares (PLS), genetic algorithm PLS (GA-PLS), principal component ANN (PC-ANN) and GA-ANN. The two proposed univariate calibration methods are, direct spectrofluorimetric method for OLM and isoabsorpitive method for determination of total concentration of OLM and AML and hence AML by subtraction. The results showed the superiority of multivariate calibration methods over univariate ones for the analysis of the binary mixture. The optimum assay conditions were established and the proposed multivariate calibration methods were successfully applied for the assay of the two drugs in validation set and combined pharmaceutical tablets with excellent recoveries. No interference was observed from common pharmaceutical additives. The results were favorably compared with those obtained by a reference spectrophotometric method. PMID:22895851

  3. A method for small-animal PET/CT alignment calibration

    NASA Astrophysics Data System (ADS)

    Pascau, J.; Vaquero, J. J.; Chamorro-Servent, J.; Rodríguez-Ruano, A.; Desco, M.

    2012-06-01

    Small-animal positron-emission tomography/computed tomography (PET/CT) scanners provide anatomical and molecular imaging, which enables the joint visualization and analysis of both types of data. A proper alignment calibration procedure is essential for small-animal imaging since resolution is much higher than that in human devices. This work presents an alignment phantom and two different calibration methods that provide a reliable and repeatable measurement of the spatial geometrical alignment between the PET and the CT subsystems of a hybrid scanner. The phantom can be built using laboratory materials, and it is meant to estimate the rigid spatial transformation that aligns both modalities. It consists of three glass capillaries filled with a positron-emitter solution and positioned in a non-coplanar triangular geometry inside the system field of view. The calibration methods proposed are both based on automatic line detection, but with different approaches to calculate the transformation of the lines between both modalities. Our results show an average accuracy of the alignment estimation of 0.39 mm over the whole field of view.

  4. Research on the method of establishing the total radiation meter calibration device

    NASA Astrophysics Data System (ADS)

    Gao, Jianqiang; Xia, Ming; Xia, Junwen; Zhang, Dong

    2015-10-01

    Pyranometer is an instrument used to measure the solar radiation, according to pyranometer differs as installation state, can be respectively measured total solar radiation, reflected radiation, or with the help of shading device for measuring scattering radiation. Pyranometer uses the principle of thermoelectric effect, inductive element adopts winding plating type multi junction thermopile, its surface is coated with black coating with high absorption rate. Hot junction in the induction surface, while the cold junction is located in the body, the cold and hot junction produce thermoelectric potential. In the linear range, the output signal is proportional to the solar irradiance. Traceability to national meteorological station, as the unit of the national legal metrology organizations, the responsibility is to transfer value of the sun and the earth radiation value about the national meteorological industry. Using the method of comparison, with indoor calibration of solar simulator, at the same location, standard pyranometer and measured pyranometer were alternately measured radiation irradiance, depending on the irradiation sensitivity standard pyranometer were calculated the radiation sensitivity of measured pyranometer. This paper is mainly about the design and calibration method of the pyranometer indoor device. The uncertainty of the calibration result is also evaluated.

  5. Preparation of Calibration Gas Mixtures Using a Dynamic Volumetric Method for Environmental Monitoring

    NASA Astrophysics Data System (ADS)

    Lee, S.; Lee, W.; Oh, S.

    2008-12-01

    It is difficult to produce and maintain the quality of calibration gas mixtures stable and reliable in gas cylinders for the identification and quantification of trace environmental contaminants. Dynamic method has attracted an attention because immediate and on-site generation of calibration gas mixture is achievable in the range of percent, sub-ppm, and micro-mole fraction. The dynamic method for the preparation of standard reference gas/vapour mixtures was carried out using polymeric permeation membrane. Permeation behaviour of various organic compounds including aliphatic and aromatic hydrocarbons and chlorohydrocarbons was tested through semi-permeable silicone polymer and fluoropolymers. Experimental results showed the constant permeation rate of target compounds under the constant temperature condition. The permeation rate, however, was largely influenced by chemical components, membrane materials, and temperature change. Mole fractions of generated vapour were calculated using the mass loss of permeation tubes and mass-flow rate of carrier gas. Dynamic volumetric techniques can be applicable for the generation of standard reference materials for the calibration of various types of organics (volatile organic compounds, polycyclic aromatic hydrocarbons, etc) in ambient air and water. This technology is also applicable to develop novel materials for passive samplers for long-term environmental monitoring.

  6. A calibration method for lateral forces for use with colloidal probe force microscopy cantilevers

    SciTech Connect

    Quintanilla, M. A. S.; Goddard, D. T.

    2008-02-15

    A calibration method is described for colloidal probe cantilevers that enables friction force measurements obtained using lateral force microscopy (LFM) to be quantified. The method is an adaptation of the lever method of Feiler et al. [A. Feiler, P. Attard, and I. Larson, Rev. Sci. Instum. 71, 2746 (2000)] and uses the advantageous positioning of probe particles that are usually offset from the central axis of the cantilever. The main sources of error in the calibration method are assessed, in particular, the potential misalignment of the long axis of the cantilever that ideally should be perpendicular to the photodiode detector. When this is not taken into account, the misalignment is shown to have a significant effect on the cantilever torsional stiffness but not on the lateral photodiode sensitivity. Also, because the friction signal is affected by the topography of the substrate, the method presented is valid only against flat substrates. Two types of particles, 20 {mu}m glass beads and UO{sub 3} agglomerates attached to silicon tapping mode cantilevers were used to test the method against substrates including glass, cleaved mica, and UO{sub 2} single crystals. Comparisons with the lateral compliance method of Cain et al. [R. G. Cain, S. Biggs, and N. W. Page, J. Colloid Interface Sci. 227, 55 (2000)] are also made.

  7. The development of in-situ calibration method for divertor IR thermography in ITER

    SciTech Connect

    Takeuchi, M.; Sugie, T.; Ogawa, H.; Takeyama, S.; Itami, K.

    2014-08-21

    For the development of the calibration method of the emissivity in IR light on the divertor plate in ITER divertor IR thermography system, the laboratory experiments have been performed by using IR instruments. The calibration of the IR camera was performed by the plane black body in the temperature of 100–600 degC. The radiances of the tungsten heated by 280 degC were measured by the IR camera without filter (2.5–5.1 μm) and with filter (2.95 μm, 4.67 μm). The preliminary data of the scattered light of the laser of 3.34 μm that injected into the tungsten were acquired.

  8. Comparison of flume and towing methods for verifying the calibration of a suspended-sediment sampler

    USGS Publications Warehouse

    Beverage, J.P.; Futrell, J.C.

    1986-01-01

    Suspended-sediment samplers must sample isokinetically (at stream velocity) in order to collect representative water samples of rivers. Each sampler solo by the Federal Interagency Sedimentation Project or by the U.S. Geological Survey Hydrologic Instrumentation Facility has been adjusted to sample isokinetically and tested in a flume to verify the calibration. The test program for a modified U.S. P-61 sampler provided an opportunity to compare flume and towing tank tests. Although the two tests yielded statistically distinct results, the difference between them was quite small. The conclusion is that verifying the calibration of any suspended-sediment sampler by either the flume or towing method should give acceptable results.

  9. Heterodyne interferometry method for calibration of a Soleil-Babinet compensator.

    PubMed

    Zhang, Wenjing; Zhang, Zhiwei

    2016-05-20

    A method based on the common-path heterodyne interferometer system is proposed for the calibration of a Soleil-Babinet compensator. In this heterodyne interferometer system, which consists of two acousto-optic modulators, the compensator being calibrated is inserted into the signal path. By using the reference beam as the benchmark and a lock-in amplifier (SR844) as the phase retardation collector, retardations of 0 and λ (one wavelength) can be located accurately, and an arbitrary retardation between 0 and λ can also be measured accurately and continuously. By fitting a straight line to the experimental data, we obtained a linear correlation coefficient (R) of 0.995, which indicates that this system is capable of linear phase detection. The experimental results demonstrate determination accuracies of 0.212° and 0.26° and measurement precisions of 0.054° and 0.608° for retardations of 0 and λ, respectively. PMID:27411154

  10. The development of in-situ calibration method for divertor IR thermography in ITER

    NASA Astrophysics Data System (ADS)

    Takeuchi, M.; Sugie, T.; Ogawa, H.; Takeyama, S.; Itami, K.

    2014-08-01

    For the development of the calibration method of the emissivity in IR light on the divertor plate in ITER divertor IR thermography system, the laboratory experiments have been performed by using IR instruments. The calibration of the IR camera was performed by the plane black body in the temperature of 100-600 degC. The radiances of the tungsten heated by 280 degC were measured by the IR camera without filter (2.5-5.1 μm) and with filter (2.95 μm, 4.67 μm). The preliminary data of the scattered light of the laser of 3.34 μm that injected into the tungsten were acquired.

  11. Note: A calibration method to determine the lumped-circuit parameters of a magnetic probe

    NASA Astrophysics Data System (ADS)

    Li, Fuming; Chen, Zhipeng; Zhu, Lizhi; Liu, Hai; Wang, Zhijiang; Zhuang, Ge

    2016-06-01

    This paper describes a novel method to determine the lumped-circuit parameters of a magnetic inductive probe for calibration by using Helmholtz coils with high frequency power supply (frequency range: 10 kHz-400 kHz). The whole calibration circuit system can be separated into two parts: "generator" circuit and "receiver" circuit. By implementing the Fourier transform, two analytical lumped-circuit models, with respect to these separated circuits, are constructed to obtain the transfer function between each other. Herein, the precise lumped-circuit parameters (including the resistance, inductance, and capacitance) of the magnetic probe can be determined by fitting the experimental data to the transfer function. Regarding the fitting results, the finite impedance of magnetic probe can be used to analyze the transmission of a high-frequency signal between magnetic probes, cables, and acquisition system.

  12. A method for automating calibration and records management for instrumentation and dosimetry

    SciTech Connect

    O`Brien, J.M. Jr.; Rushton, R.O.; Burns, R.E. Jr.

    1993-12-31

    Current industry requirements are becoming more stringent on quality assurance records and documentation for calibration of instruments and dosimetry. A novel method is presented here that will allow a progressive automation scheme to be used in pursuit of that goal. This concept is based on computer-controlled irradiators that can act as stand-alone devices or be interfaced to other components via a computer local area network. In this way, complete systems can be built with modules to create a records management system to meet the needs of small laboratories or large multi-building calibration groups. Different database engines or formats can be used simply by replacing a module. Modules for temperature and pressure monitoring or shipping and receiving can be added, as well as equipment modules for direct IEEE-488 interface to electrometers and other instrumentation.

  13. On-orbit calibration approach for star cameras based on the iteration method with variable weights.

    PubMed

    Wang, Mi; Cheng, Yufeng; Yang, Bo; Chen, Xiao

    2015-07-20

    To perform efficient on-orbit calibration for star cameras, we developed an attitude-independent calibration approach for global optimization and noise removal by least-square estimation using multiple star images, with which the optimal principal point, focal length, and the high-order focal plane distortion can be obtained in one step in full consideration of the interaction among star camera parameters. To avoid the problem when stars could be misidentified in star images, an iteration method with variable weights is introduced to eliminate the influence of misidentified star pairs. The approach can increase the precision of least-square estimation and use fewer star images. The proposed approach has been well verified to be precise and robust in three experiments. PMID:26367824

  14. Absolute calibration method for fast-streaked, fiber optic light collection, spectroscopy systems.

    SciTech Connect

    Johnston, Mark D.; Frogget, Brent; Oliver, Bryan Velten; Maron, Yitzhak; Droemer, Darryl W.; Crain, Marlon D.

    2010-04-01

    This report outlines a convenient method to calibrate fast (<1ns resolution) streaked, fiber optic light collection, spectroscopy systems. Such a system is used to collect spectral data on plasmas generated in the A-K gap of electron beam diodes fielded on the RITS-6 accelerator (8-12MV, 140-200kA). On RITS, light is collected through a small diameter (200 micron) optical fiber and recorded on a fast streak camera at the output of 1 meter Czerny-Turner monochromator (F/7 optics). To calibrate such a system, it is necessary to efficiently couple light from a spectral lamp into a 200 micron diameter fiber, split it into its spectral components, with 10 Angstroms or less resolution, and record it on a streak camera with 1ns or less temporal resolution.

  15. Neutron monitoring systems including gamma thermometers and methods of calibrating nuclear instruments using gamma thermometers

    SciTech Connect

    Moen, Stephan Craig; Meyers, Craig Glenn; Petzen, John Alexander; Foard, Adam Muhling

    2012-08-07

    A method of calibrating a nuclear instrument using a gamma thermometer may include: measuring, in the instrument, local neutron flux; generating, from the instrument, a first signal proportional to the neutron flux; measuring, in the gamma thermometer, local gamma flux; generating, from the gamma thermometer, a second signal proportional to the gamma flux; compensating the second signal; and calibrating a gain of the instrument based on the compensated second signal. Compensating the second signal may include: calculating selected yield fractions for specific groups of delayed gamma sources; calculating time constants for the specific groups; calculating a third signal that corresponds to delayed local gamma flux based on the selected yield fractions and time constants; and calculating the compensated second signal by subtracting the third signal from the second signal. The specific groups may have decay time constants greater than 5.times.10.sup.-1 seconds and less than 5.times.10.sup.5 seconds.

  16. Easy and direct method for calibrating atomic force microscopy lateral force measurements

    PubMed Central

    Liu, Wenhua; Bonin, Keith; Guthold, Martin

    2010-01-01

    We have designed and tested a new, inexpensive, easy-to-make and easy-to-use calibration standard for atomic force microscopy (AFM) lateral force measurements. This new standard simply consists of a small glass fiber of known dimensions and Young’s modulus, which is fixed at one end to a substrate and which can be bent laterally with the AFM tip at the other end. This standard has equal or less error than the commonly used method of using beam mechanics to determine a cantilever’s lateral force constant. It is transferable, thus providing a universal tool for comparing the calibrations of different instruments. It does not require knowledge of the cantilever dimensions and composition or its tip height. This standard also allows direct conversion of the photodiode signal to force and, thus, circumvents the requirement for a sensor response (sensitivity) measurement. PMID:17614616

  17. Towards in-vivo K-edge imaging using a new semi-analytical calibration method

    NASA Astrophysics Data System (ADS)

    Schirra, Carsten; Thran, Axel; Daerr, Heiner; Roessl, Ewald; Proksa, Roland

    2014-03-01

    Flat field calibration methods are commonly used in computed tomography (CT) to correct for system imperfections. Unfortunately, they cannot be applied in energy-resolving CT when using bow-tie filters owing to spectral distortions imprinted by the filter. This work presents a novel semi-analytical calibration method for photon-counting spectral CT systems, which is applicable with a bow-tie filter in place and efficiently compensates pile-up effects at fourfold increased photon flux compared to a previously published method without degradation of image quality. The achieved reduction of the scan time enabled the first K-edge imaging in-vivo. The method employs a calibration measurement with a set of flat sheets of only a single absorber material and utilizes an analytical model to predict the expected photon counts, taking into account factors such as x-ray spectrum and detector response. From the ratios of the measured x-ray intensities and the corresponding simulated photon counts, a look-up table is generated. By use of this look-up table, measured photon-counts can be corrected yielding data in line with the analytical model. The corrected data show low pixel-to-pixel variations and pile-up effects are mitigated. Consequently, operations like material decomposition based on the same analytical model yield accurate results. The method was validated on a experimental spectral CT system equipped with a bow-tie filter in a phantom experiment and an in-vivo animal study. The level of artifacts in the resulting images is considerably lower than in images generated with a previously published method. First in-vivo K-edge images of a rabbit selectively depict vessel occlusion by an ytterbium-based thermoresponsive polymer.

  18. Additives and method for controlling clathrate hydrates in fluid systems

    DOEpatents

    Sloan, E.D. Jr.; Christiansen, R.L.; Lederhos, J.P.; Long, J.P.; Panchalingam, V.; Du, Y.; Sum, A.K.W.

    1997-06-17

    Discussed is a process for preventing clathrate hydrate masses from detrimentally impeding the possible flow of a fluid susceptible to clathrate hydrate formation. The process is particularly useful in the natural gas and petroleum production, transportation and processing industry where gas hydrate formation can cause serious problems. Additives preferably contain one or more five member, six member and/or seven member cyclic chemical groupings. Additives include polymers having lactam rings. Additives can also contain polyelectrolytes that are believed to improve conformance of polymer additives through steric hindrance and/or charge repulsion. Also, polymers having an amide on which a C{sub 1}-C{sub 4} group is attached to the nitrogen and/or the carbonyl carbon of the amide may be used alone, or in combination with ring-containing polymers for enhanced effectiveness. Polymers having at least some repeating units representative of polymerizing at least one of an oxazoline, an N-substituted acrylamide and an N-vinyl alkyl amide are preferred.

  19. Additives and method for controlling clathrate hydrates in fluid systems

    DOEpatents

    Sloan, Jr., Earle Dendy; Christiansen, Richard Lee; Lederhos, Joseph P.; Long, Jin Ping; Panchalingam, Vaithilingam; Du, Yahe; Sum, Amadeu Kun Wan

    1997-01-01

    Discussed is a process for preventing clathrate hydrate masses from detrimentally impeding the possible flow of a fluid susceptible to clathrate hydrate formation. The process is particularly useful in the natural gas and petroleum production, transportation and processing industry where gas hydrate formation can cause serious problems. Additives preferably contain one or more five member, six member and/or seven member cyclic chemical groupings. Additives include polymers having lactam rings. Additives can also contain polyelectrolytes that are believed to improve conformance of polymer additives through steric hinderance and/or charge repulsion. Also, polymers having an amide on which a C.sub.1 -C.sub.4 group is attached to the nitrogen and/or the carbonyl carbon of the amide may be used alone, or in combination with ring-containing polymers for enhanced effectiveness. Polymers having at least some repeating units representative of polymerizing at least one of an oxazoline, an N-substituted acrylamide and an N-vinyl alkyl amide are preferred.

  20. A time-delay calibrated method for cornea hysteresis and intraocular pressure measurement

    NASA Astrophysics Data System (ADS)

    Wang, Kuo-Jen; Tsai, Che-Liang; Wang, Wai; Hsu, Long; Hsu, Ken-Yuh

    2016-04-01

    The presence of cornea hysteresis (CH) in characterizing the intraocular pressure (IOP) of a human eye deteriorates the accuracy of IOP. To suppress CH, the pressure gauge of a tonometer must be located as close as possible to the cornea. However, this arrangement is unpractical because appropriate working distance to the cornea is required. In this paper, a time-delay calibrated (TDC) method is proposed to counteract the undesired effect of CH in characterizing the IOP. Employing this TDC method, the CH approaches to zero for most eyes measured.

  1. A Rapid Coordinate Transformation Method Applied in Industrial Robot Calibration Based on Characteristic Line Coincidence.

    PubMed

    Liu, Bailing; Zhang, Fumin; Qu, Xinghua; Shi, Xiaojia

    2016-01-01

    Coordinate transformation plays an indispensable role in industrial measurements, including photogrammetry, geodesy, laser 3-D measurement and robotics. The widely applied methods of coordinate transformation are generally based on solving the equations of point clouds. Despite the high accuracy, this might result in no solution due to the use of ill conditioned matrices. In this paper, a novel coordinate transformation method is proposed, not based on the equation solution but based on the geometric transformation. We construct characteristic lines to represent the coordinate systems. According to the space geometry relation, the characteristic line scan is made to coincide by a series of rotations and translations. The transformation matrix can be obtained using matrix transformation theory. Experiments are designed to compare the proposed method with other methods. The results show that the proposed method has the same high accuracy, but the operation is more convenient and flexible. A multi-sensor combined measurement system is also presented to improve the position accuracy of a robot with the calibration of the robot kinematic parameters. Experimental verification shows that the position accuracy of robot manipulator is improved by 45.8% with the proposed method and robot calibration. PMID:26901203

  2. A Rapid Coordinate Transformation Method Applied in Industrial Robot Calibration Based on Characteristic Line Coincidence

    PubMed Central

    Liu, Bailing; Zhang, Fumin; Qu, Xinghua; Shi, Xiaojia

    2016-01-01

    Coordinate transformation plays an indispensable role in industrial measurements, including photogrammetry, geodesy, laser 3-D measurement and robotics. The widely applied methods of coordinate transformation are generally based on solving the equations of point clouds. Despite the high accuracy, this might result in no solution due to the use of ill conditioned matrices. In this paper, a novel coordinate transformation method is proposed, not based on the equation solution but based on the geometric transformation. We construct characteristic lines to represent the coordinate systems. According to the space geometry relation, the characteristic line scan is made to coincide by a series of rotations and translations. The transformation matrix can be obtained using matrix transformation theory. Experiments are designed to compare the proposed method with other methods. The results show that the proposed method has the same high accuracy, but the operation is more convenient and flexible. A multi-sensor combined measurement system is also presented to improve the position accuracy of a robot with the calibration of the robot kinematic parameters. Experimental verification shows that the position accuracy of robot manipulator is improved by 45.8% with the proposed method and robot calibration. PMID:26901203

  3. Investigating temporal field sampling strategies for site-specific calibration of three soil moisture - neutron intensity parameterisation methods

    NASA Astrophysics Data System (ADS)

    Iwema, J.; Rosolem, R.; Baatz, R.; Wagener, T.; Bogena, H. R.

    2015-02-01

    The Cosmic-Ray Neutron Sensor (CRNS) can provide soil moisture information at scales relevant to hydrometeorological modeling applications. Site-specific calibration is needed to translate CRNS neutron intensities into sensor footprint average soil moisture contents. We investigated temporal sampling strategies for calibration of three CRNS parameterisations (modified N0, HMF, and COSMIC) by assessing the effects of the number of sampling days and soil wetness conditions on the performance of the calibration results, for three sites with distinct climate and land use: a semi-arid site, a temperate grassland and a temperate forest. When calibrated with a year of data, COSMIC performed relatively good at all three sites, and the modified N0 method performed best at the two humid sites. It is advisable to collect soil moisture samples on more than a single day regardless of which parameterisation is used. In any case, sampling on more than ten days would, despite the strong increase in work effort, improve calibration results only little. COSMIC needed the least number of days at each site. At the semi-arid site, the N0mod method was calibrated better under average wetness conditions, whereas HMF and COSMIC were calibrated better under drier conditions. Average soil wetness condition gave better calibration results at the two humid sites. The calibration results for the HMF method were better when calibrated with combinations of days with similar soil wetness conditions, opposed to N0mod and COSMIC, which profited from using days with distinct wetness conditions. The outcomes of this study can be used by researchers as a CRNS calibration strategy guideline.

  4. Accurate and quick calibration method for polarization-modulation spectroscopy using an ac-modulated polarizing undulator

    SciTech Connect

    Tanaka, Masahito; Yagi-Watanabe, Kazutoshi; Kaneko, Fusae; Nakagawa, Kazumichi

    2008-08-15

    An accurate calibration method in which an ac-modulated polarizing undulator is used for polarization modulation spectroscopy such as circular dichroism (CD) and linear dichroism (LD) has been proposed and successfully applied to vacuum ultraviolet (vuv) CD and LD spectra measured at beamline BL-5B in the electron storage ring, TERAS, at AIST. This calibration method employs an undulator-modulation spectroscopic method with a multireflection polarimeter, and it uses electronic and optical elements identical to those used for the CD and LD measurements. This method regards the polarimeter as a standard sample for the CD and LD measurements in the vuv region in which a standard sample has not yet been established. The calibration factors for the CD and LD spectra are obtained over a wide range of wavelengths, from 120 to 230 nm, at TERAS BL-5B. The calibrated CD and LD spectra measured at TERAS exhibit good agreement with the standard spectra for wavelengths greater than 170 nm; the mean differences between the standard and calibrated CD and LD spectra are approximately 7% and 4%, respectively. This method enables a remarkable reduction in the experimental time, from approximately 1 h to less than 10 min that is sufficient to observe the storage-ring current dependence of the calibration factors. This method can be applied to the calibration of vuv-CD spectra measured using a conventional photoelastic modulator and for performing an accurate analysis of protein secondary structures.

  5. Performance of three reflectance calibration methods for airborne hyperspectral spectrometer data.

    PubMed

    Miura, Tomoaki; Huete, Alfredo R

    2009-01-01

    In this study, the performances and accuracies of three methods for converting airborne hyperspectral spectrometer data to reflectance factors were characterized and compared. The "reflectance mode (RM)" method, which calibrates a spectrometer against a white reference panel prior to mounting on an aircraft, resulted in spectral reflectance retrievals that were biased and distorted. The magnitudes of these bias errors and distortions varied significantly, depending on time of day and length of the flight campaign. The "linear-interpolation (LI)" method, which converts airborne spectrometer data by taking a ratio of linearly-interpolated reference values from the preflight and post-flight reference panel readings, resulted in precise, but inaccurate reflectance retrievals. These reflectance spectra were not distorted, but were subject to bias errors of varying magnitudes dependent on the flight duration length. The "continuous panel (CP)" method uses a multi-band radiometer to obtain continuous measurements over a reference panel throughout the flight campaign, in order to adjust the magnitudes of the linear-interpolated reference values from the preflight and post-flight reference panel readings. Airborne hyperspectral reflectance retrievals obtained using this method were found to be the most accurate and reliable reflectance calibration method. The performances of the CP method in retrieving accurate reflectance factors were consistent throughout time of day and for various flight durations. Based on the dataset analyzed in this study, the uncertainty of the CP method has been estimated to be 0.0025 ± 0.0005 reflectance units for the wavelength regions not affected by atmospheric absorptions. The RM method can produce reasonable results only for a very short-term flight (e.g., < 15 minutes) conducted around a local solar noon. The flight duration should be kept shorter than 30 minutes for the LI method to produce results with reasonable accuracies. An important

  6. Performance of Three Reflectance Calibration Methods for Airborne Hyperspectral Spectrometer Data

    PubMed Central

    Miura, Tomoaki; Huete, Alfredo R.

    2009-01-01

    In this study, the performances and accuracies of three methods for converting airborne hyperspectral spectrometer data to reflectance factors were characterized and compared. The “reflectance mode (RM)” method, which calibrates a spectrometer against a white reference panel prior to mounting on an aircraft, resulted in spectral reflectance retrievals that were biased and distorted. The magnitudes of these bias errors and distortions varied significantly, depending on time of day and length of the flight campaign. The “linear-interpolation (LI)” method, which converts airborne spectrometer data by taking a ratio of linearly-interpolated reference values from the preflight and post-flight reference panel readings, resulted in precise, but inaccurate reflectance retrievals. These reflectance spectra were not distorted, but were subject to bias errors of varying magnitudes dependent on the flight duration length. The “continuous panel (CP)” method uses a multi-band radiometer to obtain continuous measurements over a reference panel throughout the flight campaign, in order to adjust the magnitudes of the linear-interpolated reference values from the preflight and post-flight reference panel readings. Airborne hyperspectral reflectance retrievals obtained using this method were found to be the most accurate and reliable reflectance calibration method. The performances of the CP method in retrieving accurate reflectance factors were consistent throughout time of day and for various flight durations. Based on the dataset analyzed in this study, the uncertainty of the CP method has been estimated to be 0.0025 ± 0.0005 reflectance units for the wavelength regions not affected by atmospheric absorptions. The RM method can produce reasonable results only for a very short-term flight (e.g., < 15 minutes) conducted around a local solar noon. The flight duration should be kept shorter than 30 minutes for the LI method to produce results with reasonable accuracies

  7. A calibration transmission method to determine the gamma-ray linear attenuation coefficient without a collimator.

    PubMed

    Byun, Jong-In; Yun, Ju-Yong

    2015-08-01

    It is shown that the gamma-ray linear attenuation coefficient of a sample with unknown chemical composition can be determined through a systematic calibration of the correlation between the linear attenuation coefficient, gamma-ray energy and the relative degree of attenuation. For calibration, H2O, MnO2, NaCl, Na2CO3 and (NH4)2SO4 were used as reference materials. Point-like gamma-ray sources with modest activity of approximately 37kBq, along with an HPGe detector, were used in the measurements. A semi-empirical formula was derived to calculate the linear attenuation coefficients as a function of the relative count rate and the gamma-ray energy. The method was applied to the determination of the linear attenuation coefficients for K2CrO4 and SiO2 test samples in the same setup used in calibration. The experimental result agreed well with the ones calculated by elementary data. PMID:25997111

  8. A calibration method for patient specific IMRT QA using a single therapy verification film

    PubMed Central

    Shukla, Arvind Kumar; Oinam, Arun S.; Kumar, Sanjeev; Sandhu, I.S.; Sharma, S.C.

    2013-01-01

    Aim The aim of the present study is to develop and verify the single film calibration procedure used in intensity-modulated radiation therapy (IMRT) quality assurance. Background Radiographic films have been regularly used in routine commissioning of treatment modalities and verification of treatment planning system (TPS). The radiation dosimetery based on radiographic films has ability to give absolute two-dimension dose distribution and prefer for the IMRT quality assurance. However, the single therapy verification film gives a quick and significant reliable method for IMRT verification. Materials and methods A single extended dose rate (EDR 2) film was used to generate the sensitometric curve of film optical density and radiation dose. EDR 2 film was exposed with nine 6 cm × 6 cm fields of 6 MV photon beam obtained from a medical linear accelerator at 5-cm depth in solid water phantom. The nine regions of single film were exposed with radiation doses raging from 10 to 362 cGy. The actual dose measurements inside the field regions were performed using 0.6 cm3 ionization chamber. The exposed film was processed after irradiation using a VIDAR film scanner and the value of optical density was noted for each region. Ten IMRT plans of head and neck carcinoma were used for verification using a dynamic IMRT technique, and evaluated using the gamma index method with TPS calculated dose distribution. Results Sensitometric curve has been generated using a single film exposed at nine field region to check quantitative dose verifications of IMRT treatments. The radiation scattered factor was observed to decrease exponentially with the increase in the distance from the centre of each field region. The IMRT plans based on calibration curve were verified using the gamma index method and found to be within acceptable criteria. Conclusion The single film method proved to be superior to the traditional calibration method and produce fast daily film calibration for highly

  9. A Method to Solve Interior and Exterior Camera Calibration Parameters for Image Resection

    NASA Technical Reports Server (NTRS)

    Samtaney, Ravi

    1999-01-01

    An iterative method is presented to solve the internal and external camera calibration parameters, given model target points and their images from one or more camera locations. The direct linear transform formulation was used to obtain a guess for the iterative method, and herein lies one of the strengths of the present method. In all test cases, the method converged to the correct solution. In general, an overdetermined system of nonlinear equations is solved in the least-squares sense. The iterative method presented is based on Newton-Raphson for solving systems of nonlinear algebraic equations. The Jacobian is analytically derived and the pseudo-inverse of the Jacobian is obtained by singular value decomposition.

  10. Symmetrizable connection and combined calibration method for accuracy measurement of CMM

    NASA Astrophysics Data System (ADS)

    Fei, Yetai; Xie, Shao-Feng; Chen, Xia-Huai

    1993-09-01

    In this paper, a new method of syrmetrizab1e connection and ciribined calibration is presented based on an analysis of accuracy of the 1Mt The novel measuring principle and succinct mathematical model are described. By experimental ctparison, the correctness and practicability of the method are proved. In order to assess the M,i accuracy and conpensate for errors, all errors should be measured with high accuracy and efficiency. At the same time, a succinct mathematical model should be developed. For this reason seeking an efficient measuring method for the IM1 has all along been an important subject in this field. All currently used measuring methods for the lMv! have their limitations. To remedy such a situation, the synmetrizable connection method is presented. It solves current problems of the M1 accuracy verification.

  11. An inexpensive method for kinematic calibration of a parallel robot by using one hand-held camera as main sensor.

    PubMed

    Traslosheros, Alberto; Sebastián, José María; Torrijos, Jesús; Carelli, Ricardo; Castillo, Eduardo

    2013-01-01

    This paper presents a novel method for the calibration of a parallel robot, which allows a more accurate configuration instead of a configuration based on nominal parameters. It is used, as the main sensor with one camera installed in the robot hand that determines the relative position of the robot with respect to a spherical object fixed in the working area of the robot. The positions of the end effector are related to the incremental positions of resolvers of the robot motors. A kinematic model of the robot is used to find a new group of parameters, which minimizes errors in the kinematic equations. Additionally, properties of the spherical object and intrinsic camera parameters are utilized to model the projection of the object in the image and thereby improve spatial measurements. Finally, several working tests, static and tracking tests are executed in order to verify how the robotic system behaviour improves by using calibrated parameters against nominal parameters. In order to emphasize that, this proposed new method uses neither external nor expensive sensor. That is why new robots are useful in teaching and research activities. PMID:23921827

  12. System and method for high power diode based additive manufacturing

    DOEpatents

    El-Dasher, Bassem S.; Bayramian, Andrew; Demuth, James A.; Farmer, Joseph C.; Torres, Sharon G.

    2016-04-12

    A system is disclosed for performing an Additive Manufacturing (AM) fabrication process on a powdered material forming a substrate. The system may make use of a diode array for generating an optical signal sufficient to melt a powdered material of the substrate. A mask may be used for preventing a first predetermined portion of the optical signal from reaching the substrate, while allowing a second predetermined portion to reach the substrate. At least one processor may be used for controlling an output of the diode array.

  13. An improved method for calibrating time-of-flight Laue single-crystal neutron diffractometers

    PubMed Central

    Bull, Craig L.; Johnson, Michael W.; Hamidov, Hayrullo; Komatsu, Kazuki; Guthrie, Malcolm; Gutmann, Matthias J.; Loveday, John S.; Nelmes, Richard J.

    2014-01-01

    A robust and comprehensive method for determining the orientation matrix of a single-crystal sample using the neutron Laue time-of-flight (TOF) technique is described. The new method enables the measurement of the unit-cell parameters with an uncertainty in the range 0.015–0.06%, depending upon the crystal symmetry and the number of reflections measured. The improved technique also facilitates the location and integration of weak reflections, which are often more difficult to discern amongst the increased background at higher energies. The technique uses a mathematical model of the relative positions of all the detector pixels of the instrument, together with a methodology that establishes a reproducible reference frame and a method for determining the parameters of the instrument detector model. Since all neutron TOF instruments require precise detector calibration for their effective use, it is possible that the method described here may be of use on other instruments where the detector calibration cannot be determined by other means. PMID:24904244

  14. Calibration independent generalized cavity method for microwave characterization of powdered materials

    NASA Astrophysics Data System (ADS)

    Jha, Abhishek Kumar; Azizurrahaman; Akhtar, Mohammad Jaleel

    2015-06-01

    A generalized cavity method relaxing the major assumptions of conventional cavity perturbation technique is presented for characterization of the powdered sample in the microwave frequency range. The unified method, which is based on the inverse optimization technique, eliminates the complexity of measurement caused due to the existence of sample holder and produces an accurate result. In this paper, an attractive numerical calibration approach is proposed in lieu of the practical calibration technique which usually requires either a set of standards or a number of reference samples. The sample holder especially made of borosilicate glass is designed to contain the powdered samples, and the X-band rectangular cavity is fabricated. For verification of the proposed technique, the pulverized alumina and polyethylene oxide with various packing fractions are measured using the fabricated cavity and the vector network analyzer. The dielectric constant of these samples is extracted using the proposed unified approach which is found to be in good agreement with the theoretical data obtained by Landau-Lifshitz and Looyenga model. The accuracy of the proposed generalized cavity method for powdered samples is found to be better than the conventional resonator methods available in the literature.

  15. Calibration independent generalized cavity method for microwave characterization of powdered materials.

    PubMed

    Jha, Abhishek Kumar; Azizurrahaman; Akhtar, Mohammad Jaleel

    2015-06-01

    A generalized cavity method relaxing the major assumptions of conventional cavity perturbation technique is presented for characterization of the powdered sample in the microwave frequency range. The unified method, which is based on the inverse optimization technique, eliminates the complexity of measurement caused due to the existence of sample holder and produces an accurate result. In this paper, an attractive numerical calibration approach is proposed in lieu of the practical calibration technique which usually requires either a set of standards or a number of reference samples. The sample holder especially made of borosilicate glass is designed to contain the powdered samples, and the X-band rectangular cavity is fabricated. For verification of the proposed technique, the pulverized alumina and polyethylene oxide with various packing fractions are measured using the fabricated cavity and the vector network analyzer. The dielectric constant of these samples is extracted using the proposed unified approach which is found to be in good agreement with the theoretical data obtained by Landau-Lifshitz and Looyenga model. The accuracy of the proposed generalized cavity method for powdered samples is found to be better than the conventional resonator methods available in the literature. PMID:26144529

  16. The three-measurement two-calibration method for measuring the transfer matrix.

    PubMed

    Rodriguez, S; Gibiat, V; Lefebvre, A; Guilain, Stephane

    2011-05-01

    Extensive use of transfer matrices (TMs) is made in determining the acoustic properties of a duct and in in-duct acoustic propagation models in the automotive industry and for musical acoustics purposes. The experimental apparatuses of classical TM measurement methods feature two measurement heads. Two microphones are flush with the walls of each head. The pressure signals are processed following the transfer function method constructed on an analytical model of acoustic propagation in measurement heads. The present paper aims at presenting a measurement method based on a three-microphone experimental apparatus and on its acoustic calibration through two reference measurements: the three-measurement two-calibration method for measuring the TM (3M2C-TM). Two microphones are flush with the measurement head walls and one is in the cap closing one side of the measured duct. 3M2C-TM proved essential for an accurate measurement of the four TM elements of two different ducts: a cylindrical duct and an expansion chamber. PMID:21568409

  17. A calibration method of self-referencing interferometry based on maximum likelihood estimation

    NASA Astrophysics Data System (ADS)

    Zhang, Chen; Li, Dahai; Li, Mengyang; E, Kewei; Guo, Guangrao

    2015-05-01

    Self-referencing interferometry has been widely used in wavefront sensing. However, currently the results of wavefront measurement include two parts, one is the real phase information of wavefront under test and the other is the system error in self-referencing interferometer. In this paper, a method based on maximum likelihood estimation is presented to calibrate the system error in self-referencing interferometer. Firstly, at least three phase difference distributions are obtained by three position measurements of the tested component: one basic position, one rotation and one lateral translation. Then, combining the three phase difference data and using the maximum likelihood method to create a maximum likelihood function, reconstructing the wavefront under test and the system errors by least square estimation and Zernike polynomials. The simulation results show that the proposed method can deal with the issue of calibration of a self-referencing interferometer. The method can be used to reduce the effect of system errors on extracting and reconstructing the wavefront under test, and improve the measurement accuracy of the self-referencing interferometer.

  18. Cloned plasmid DNA fragments as calibrators for controlling GMOs: different real-time duplex quantitative PCR methods.

    PubMed

    Taverniers, Isabel; Van Bockstaele, Erik; De Loose, Marc

    2004-03-01

    Analytical real-time PCR technology is a powerful tool for implementation of the GMO labeling regulations enforced in the EU. The quality of analytical measurement data obtained by quantitative real-time PCR depends on the correct use of calibrator and reference materials (RMs). For GMO methods of analysis, the choice of appropriate RMs is currently under debate. So far, genomic DNA solutions from certified reference materials (CRMs) are most often used as calibrators for GMO quantification by means of real-time PCR. However, due to some intrinsic features of these CRMs, errors may be expected in the estimations of DNA sequence quantities. In this paper, two new real-time PCR methods are presented for Roundup Ready soybean, in which two types of plasmid DNA fragments are used as calibrators. Single-target plasmids (STPs) diluted in a background of genomic DNA were used in the first method. Multiple-target plasmids (MTPs) containing both sequences in one molecule were used as calibrators for the second method. Both methods simultaneously detect a promoter 35S sequence as GMO-specific target and a lectin gene sequence as endogenous reference target in a duplex PCR. For the estimation of relative GMO percentages both "delta C(T)" and "standard curve" approaches are tested. Delta C(T) methods are based on direct comparison of measured C(T) values of both the GMO-specific target and the endogenous target. Standard curve methods measure absolute amounts of target copies or haploid genome equivalents. A duplex delta C(T) method with STP calibrators performed at least as well as a similar method with genomic DNA calibrators from commercial CRMs. Besides this, high quality results were obtained with a standard curve method using MTP calibrators. This paper demonstrates that plasmid DNA molecules containing either one or multiple target sequences form perfect alternative calibrators for GMO quantification and are especially suitable for duplex PCR reactions. PMID:14689155

  19. NMR Stark Spectroscopy: New Methods to Calibrate NMR Sensitivity to Electric Fields

    NASA Astrophysics Data System (ADS)

    Tarasek, Matthew R.

    The influence of electrostatics on NMR parameters is well accepted. Thus, NMR is a promising route to probe electrical features within molecules and materials. However, applications of NMR Stark effects (E-field induced changes in spin energy levels) have been elusive. I have developed new approaches to resolve NMR Stark effects from an applied E field. This calibrates nuclear probes whose spectral response might later be used to evaluate internal E fields that are critical to function, such as those due to local charge distributions or sample structure. I will present two novel experimental approaches for direct calibration of NMR quadrupolar Stark effects (QSEs). In the first, steady-state (few-second) excitation by an E field at twice the NMR frequency (2ω 0) is used to saturate spin magnetization. The extent of saturation vs. E-field amplitude calibrates the QSE response rate, while measurements vs sample orientation determine tensorial character. The second method instead synchronizes short (few µs) pulses of the 2ω0 E field with a multiple-pulse NMR sequence. This, “POWER” (Perturbations Observed With Enhanced Resolution) approach enables more accurate measure of small QSEs (i.e. few Hz spectral changes). A 2nd key advantage is the ability to define tensorial response without reorienting the sample, but instead varying the phase of the 2ω0 field. I will describe these experiments and my home-built NMR “Stark probe”, employed on a conventional wide-bore solid-state NMR system. Results with GaAs demonstrate each method, while extensions to a wider array of molecular and material systems may now be possible using these methods.

  20. [Application of calibration curve method and partial least squares regression analysis to quantitative analysis of nephrite samples using XRF].

    PubMed

    Liu, Song; Su, Bo-min; Li, Qing-hui; Gan, Fu-xi

    2015-01-01

    The authors tried to find a method for quantitative analysis using pXRF without solid bulk stone/jade reference samples. 24 nephrite samples were selected, 17 samples were calibration samples and the other 7 are test samples. All the nephrite samples were analyzed by Proton induced X-ray emission spectroscopy (PIXE) quantitatively. Based on the PIXE results of calibration samples, calibration curves were created for the interested components/elements and used to analyze the test samples quantitatively; then, the qualitative spectrum of all nephrite samples were obtained by pXRF. According to the PIXE results and qualitative spectrum of calibration samples, partial least square method (PLS) was used for quantitative analysis of test samples. Finally, the results of test samples obtained by calibration method, PLS method and PIXE were compared to each other. The accuracy of calibration curve method and PLS method was estimated. The result indicates that the PLS method is the alternate method for quantitative analysis of stone/jade samples. PMID:25993858

  1. A calibration method for optical trap force by use of electrokinetic phenomena

    NASA Astrophysics Data System (ADS)

    Yu, Youli; Zhang, Zhenxi; Zhang, Xiaolin

    2006-09-01

    An experimental method for calibration of optical trap force upon cells by use of electrokinetic phenomena is demonstrated. An electronkinetic sample chamber system (ESCS) is designed instead of a common sample chamber and a costly automatism stage, thus the experimental setup is simpler and cheaper. Experiments indicate that the range of the trap force measured by this method is piconewton and sub-piconewton, which makes it fit for study on non-damage interaction between light and biological particles with optical tweezers especially. Since this method is relevant to particle electric charge, by applying an alternating electric field, the new method may overcome the problem of correcting drag force and allow us to measure simultaneously optical trap stiffness and particle electric charge.

  2. Global and Local Optimization Methods for Calibration and Design of Computationally Expensive Hydrologic Models

    NASA Astrophysics Data System (ADS)

    Shoemaker, C. A.; Singh, A.

    2008-12-01

    This paper will describe some new optimization algorithms and their application to hydrologic models. The approaches include a parallel version of a new heuristic algorithm combined with tabu search and a mathematically derived global optimization method that is based on trust region methods. The goals of these methods are to find optimal solutions to calibration problems and to design problems with relatively few simulations or (in a parallel environment) relatively little wallclock time. This is important because currently it is not possible to apply global optimization methods like genetic algorithms to computationally expensive simulation models like partial differential equations (with many nodes in groundwater) because it is not feasible to do thousands of simulations to evaluate the objective/fitness function. Results of the application of the algorithms to some complex models of groundwater contamination and phosphorous transport in watersheds will be presented.

  3. A New Method to Calibrate Attachment Angles of Data Loggers in Swimming Sharks

    NASA Astrophysics Data System (ADS)

    Kawatsu, Shizuka; Sato, Katsufumi; Watanabe, Yuuki; Hyodo, Susumu; Breves, Jason P.; Fox, Bradley K.; Grau, E. Gordon; Miyazaki, Nobuyuki

    2009-12-01

    Recently, animal-borne accelerometers have been used to record the pitch angle of aquatic animals during swimming. When evaluating pitch angle, it is necessary to consider a discrepancy between the angle of an accelerometer and the long axis of an animal. In this study, we attached accelerometers to 17 free-ranging scalloped hammerhead shark ( Sphyrna lewini) pups from Kaneohe Bay, Hawaii. Although there are methods to calibrate attachment angles of accelerometers, we confirmed that previous methods were not applicable for hammerhead pups. According to raw data, some sharks ascended with a negative angle, which differs from tank observations of captive sharks. In turn, we developed a new method to account for this discrepancy in swimming sharks by estimating the attachment angle from the relationship between vertical speed (m/s) and pitch angle obtained by each accelerometer. The new method can be utilized for field observation of a wide range of species.

  4. Towards a shock tube method for the dynamic calibration of pressure sensors

    PubMed Central

    Downes, Stephen; Knott, Andy; Robinson, Ian

    2014-01-01

    In theory, shock tubes provide a pressure change with a very fast rise time and calculable amplitude. This pressure step could provide the basis for the calibration of pressure transducers used in highly dynamic applications. However, conventional metal shock tubes can be expensive, unwieldy and difficult to modify. We describe the development of a 1.4 MPa (maximum pressure) shock tube made from unplasticized polyvinyl chloride pressure tubing which provides a low-cost, light and easily modifiable basis for establishing a method for determining the dynamic characteristics of pressure sensors. PMID:25071242

  5. Molecular branching ratio method for intensity calibration of optical systems in the vacuum ultraviolet

    NASA Technical Reports Server (NTRS)

    Mumma, M. J.

    1972-01-01

    A state-of-the-art review is given of the molecular branching ratio method for intensity calibration in the vacuum ultraviolet. Ways are described for determining both relative and quantitative responses in the wavelength range 1000 A to 3000 A. The molecular band systems which are discussed are the following: H2(B 1 Sigma u +)-(X 1 Sigma g +), H2(C 1 Pi u)-(X 1 Sigma g +), N2(A 1 Pi g)-(X 1 Sigma g +), CO(A 1 Pi)-(X 1 Sigma +), NO(A 2 Sigma +)-(X 2 Pi r), and NO(+) (A 1 Pi)-(X 1 Sigma +).

  6. Linear Calibration of Radiographic Mineral Density Using Video-Digitizing Methods

    NASA Technical Reports Server (NTRS)

    Martin, R. Bruce; Papamichos, Thomas; Dannucci, Greg A.

    1990-01-01

    Radiographic images can provide quantitative as well as qualitative information if they are subjected to densitometric analysis. Using modem video-digitizing techniques, such densitometry can be readily accomplished using relatively inexpensive computer systems. However, such analyses are made more difficult by the fact that the density values read from the radiograph have a complex, nonlinear relationship to bone mineral content. This article derives the relationship between these variables from the nature of the intermediate physical processes, and presents a simple mathematical method for obtaining a linear calibration function using a step wedge or other standard.

  7. Linear Calibration of Radiographic Mineral Density Using Video-Digitizing Methods

    NASA Technical Reports Server (NTRS)

    Martin, R. Bruce; Papamichos, Thomas; Dannucci, Greg A.

    1990-01-01

    Radiographic images can provide quantitative as well as qualitative information if they are subjected to densitometric analysis. Using modern video-digitizing techniques, such densitometry can be readily accomplished using relatively inexpensive computer systems. However, such analyses are made more difficult by the fact that the density values read from the radiograph have a complex, nonlinear relationship to bone mineral content. This article derives the relationship between these variables from the nature of the intermediate physical processes, and presents a simple mathematical method for obtaining a linear calibration function using a step wedge or other standard.

  8. Calibration Method in Elasticity Evaluation of Regenerating Cartilage Based on Ultrasonic Particle Velocity

    NASA Astrophysics Data System (ADS)

    Nitta, Naotaka; Hyodo, Koji; Misawa, Masaki; Hayashi, Kazuhiko; Shirasaki, Yoshio; Homma, Kazuhiro; Shiina, Tsuyoshi

    2013-07-01

    It is important in regenerative medicine to evaluate the maturity of regenerating tissue. In the maturity evaluation of regenerating cartilage, it is useful to measure the temporal change in elasticity because the maturity of regenerating tissue is closely related to its elasticity. In this study, a quantitative elasticity evaluation of extracted regenerating cartilage samples, which is based on the laser Doppler measurement of ultrasonic particle velocity and calibration, was experimentally investigated using agar-based phantoms with different Young's moduli and regenerating cartilage samples extracted from beagles in animal experiments. The experimental results verified the feasibility of the proposed method for the elasticity evaluation of regenerating cartilage samples.

  9. Optimization of method a load cell calibration for the measurement of coefficient of friction

    NASA Astrophysics Data System (ADS)

    Castro, R. M.; Pereira, M.; Sousa, A. R.; Curi, E. I. M.; Izidoro, C. L.; Correa, L. C.

    2016-07-01

    The instrumentation of equipment for mechanical testing is used to optimize the time to deliver a result, besides minimizing errors associated with manual measurements. Given this context, this work aims to present a calibration method for a load cell to determine the measurement results of force and friction coefficient, developed from on rotary pin-on-disk tribometer. The results indicate that the procedure provides measurements reliable for the tribological phenomena, resulting in with proximity the values provided by the ASTM G99-04.

  10. Calibration period dependence of extreme flood estimations (with a model-based flood frequency method)

    NASA Astrophysics Data System (ADS)

    Brigode, P.; Bernardara, P.; Paquet, E.; Gailhard, J.; Garavaglia, F.; Ribstein, P.; Micovic, Z.

    2013-12-01

    Extreme floods estimation methods are developed since many years within the hydrological and statistical communities. More recently, approaches based on the statistical analysis of flood streamflow samples simulated by rainfall-runoff models which are forced by simulated rainfall spread in the scientific literature. These approaches, called stochastic simulation methods, are typically composed by a probabilistic rainfall model and a rainfall-runoff model. Each of these two models are calibrated over observed hydrometeorological series such as daily precipitation series for the probabilistic rainfall models or such as daily streamflow, precipitation and temperature series for the rainfall-runoff models. Since extreme flood observations are by definition particularly rare, the validation of the proposed extreme flood estimations is one of the main critical issues, whatever the method - statistical or physically-based - used. Moreover, the observed hydrometeorological series used for the calibration of the stochastic simulation methods may be subject to significant variability over time, due to global climate oscillations such as El Niño Southern Oscillations for example. If the estimation of total involved uncertainty is a difficult task, investigating to what extent the proposed extreme flood values are dependent on the calibration period is an interesting first step. The general aim of this study is to propose a methodology for performing a sensitivity analysis of extreme flood estimations to the variability of observed series used for the model calibrations in a stochastic simulation framework. The methodology proposed is based on the nonparametric bootstrap concept and consists to perform a set of block-bootstrap experiments, thus generating different sets of observed series sub-samples. The generated observed series sub-samples are then used for the calibration of the different models considered within the stochastic simulation method. The main originality of

  11. Locally-calibrated light transmission visualization methods to quantify nonaqueous phase liquid mass in porous media

    NASA Astrophysics Data System (ADS)

    Wang, Huaguo; Chen, Xiaosong; Jawitz, James W.

    2008-11-01

    Five locally-calibrated light transmission visualization (LTV) methods were tested to quantify nonaqueous phase liquid (NAPL) mass and mass reduction in porous media. Tetrachloroethylene (PCE) was released into a two-dimensional laboratory flow chamber packed with water-saturated sand which was then flushed with a surfactant solution (2% Tween 80) until all of the PCE had been dissolved. In all the LTV methods employed here, the water phase was dyed, rather than the more common approach of dyeing the NAPL phase, such that the light adsorption characteristics of NAPL did not change as dissolution progressed. Also, none of the methods used here required the use of external calibration chambers. The five visualization approaches evaluated included three methods developed from previously published models, a binary method, and a novel multiple wavelength method that has the advantage of not requiring any assumptions about the intra-pore interface structure between the various phases (sand/water/NAPL). The new multiple wavelength method is also expected to be applicable to any translucent porous media containing two immiscible fluids (e.g., water-air, water-NAPL). Results from the sand-water-PCE system evaluated here showed that the model that assumes wetting media of uniform pore size (Model C of Niemet and Selker, 2001) and the multiple wavelength model with no interface structure assumptions were able to accurately quantify PCE mass reduction during surfactant flushing. The average mass recoveries from these two imaging methods were greater than 95% for domain-average NAPL saturations of approximately 2.6 × 10 - 2 , and were approximately 90% during seven cycles of surfactant flushing that sequentially reduced the average NAPL saturation to 7.5 × 10 - 4 .

  12. Direct liquid deposition calibration method for trace cyclotrimethylenetrinitramine using thermal desorption instrumentation.

    PubMed

    Field, Christopher R; Lubrano, Adam L; Rogers, Duane A; Giordano, Braden C; Collins, Greg E

    2013-03-22

    A simple method for establishing calibration curves with sorbent-filled thermal desorption tubes has been demonstrated for nitroaromatic and nitramine vapor samples using a thermal desorption system with a cooled inlet system (TDS-CIS), which was coupled to a gas chromatograph (GC) with an electron capture detector (ECD). The method relies upon the direct liquid deposition of standard solutions onto the glass frit at the head of sorbent-filled thermal desorption tubes. Linear calibration results and ideal system conditions for the TDS-CIS-GC-ECD were established for mixtures containing both cyclotrimethylenetrinitramine, a.k.a. RDX, and 2,4,6-trinitrotoluene (TNT). Because of the chemical characteristics of RDX, a higher TDS-CIS flow rate relative to the optimized approach for TNT was required for efficient RDX desorption. Simultaneous quantitation of TNT and RDX using the direct liquid deposition method with optimized instrumentation parameters for RDX were compared to results from a standard split/splitless GC inlet and a CIS. PMID:23415141

  13. Method to calibrate the absolute energy scale of air showers with ultrahigh energy photons.

    PubMed

    Homola, Piotr; Risse, Markus

    2014-04-18

    Calibrating the absolute energy scale of air showers initiated by ultrahigh energy (UHE) cosmic rays is an important experimental issue. Currently, the corresponding systematic uncertainty amounts to 14%-21% using the fluorescence technique. Here, we describe a new, independent method which can be applied if ultrahigh energy photons are observed. While such photon-initiated showers have not yet been identified, the capabilities of present and future cosmic-ray detectors may allow their discovery. The method makes use of the geomagnetic conversion of UHE photons (preshower effect), which significantly affects the subsequent longitudinal shower development. The conversion probability depends on photon energy and can be calculated accurately by QED. The comparison of the observed fraction of converted photon events to the expected one allows the determination of the absolute energy scale of the observed photon air showers and, thus, an energy calibration of the air shower experiment. We provide details of the method and estimate the accuracy that can be reached as a function of the number of observed photon showers. Already a very small number of UHE photons may help to test and fix the absolute energy scale. PMID:24785024

  14. Calibration-free laser induced breakdown spectroscopy as an alternative method for found meteorite fragments analysis

    NASA Astrophysics Data System (ADS)

    Horňáčková, Michaela; Plavčan, Jozef; Rakovský, Jozef; Porubčan, Vladimír; Ozdín, Daniel; Veis, Pavel

    2014-04-01

    Calibration-free laser induced breakdown spectroscopy (CF-LIBS) was used for the determination of elemental composition and quantitative analysis of the Košice meteorite by means of time resolved and broadband emission spectroscopy (200-1000 nm). The electron temperature was determined using the Saha-Boltzmann plot method and the electron density from Stark broadening of the hydrogen Hα line (656 nm). Apart from magnesium, silicon and iron, which are the main elemental constituents of examined meteorite fragments, elements such as aluminum, nickel, potassium, sodium, chromium, calcium and manganese were also identified in the obtained LIBS spectra. Concentrations of Al, Ca, Cr, Fe, Mg, Mn, Na, Ni and Si were calculated using the calibration free approach and results were compared with ones obtained from the ICP-MS analyses. For the increase of the CF-LIBS accuracy, a selection of spectral lines was performed. Considering the transition probability, the population of absorbing level, the degree of ionization and predicted elemental concentration we calculated the probability of self-absorption and, consequently, spectral lines with highest self-absorption probability were rejected. CF-LIBS can be used as an alternative method for the meteorite fragments analysis (including the inner part and crust), because this method is quasi non-destructive and therefore analysis of all found fragments with minimal destruction is possible.

  15. An Improvement of Pose Measurement Method Using Global Control Points Calibration

    PubMed Central

    Sun, Changku; Sun, Pengfei; Wang, Peng

    2015-01-01

    During the last decade pose measurement technologies have gained an increasing interest in the computer vision. The vision-based pose measurement method has been widely applied in complex environments. However, the pose measurement error is a problem in the measurement applications. It grows rapidly with increasing measurement range. In order to meet the demand of high accuracy in large measurement range, a measurement error reduction solution to the vision-based pose measurement method, called Global Control Point Calibration (GCPC), is proposed. GCPC is an optimized process of existing visual pose measurement methods. The core of GCPC is to divide the measurement error into two types: the control point error and the control space error. Then by creating the global control points as well as performing error calibration of object pose, the two errors are processed. The control point error can be eliminated and the control space error is minimized. GCPC is experimented on the moving target in the camera’s field of view. The results show that the RMS error is 0.175° in yaw angle, 0.189° in pitch angle, and 0.159° in roll angle, which demonstrate that GCPC works effectively and stably. PMID:26207825

  16. Simultaneous spectrophotometric determination of chromium(VI) and iron (III) by H-point standard addition method

    NASA Astrophysics Data System (ADS)

    Larionova, E. V.; Bulygina, K. A.

    2016-02-01

    In this work the possibility of simultaneous spectrophotometric determination of chromium (VI) and iron (III) in alloys with help of the mixed organic reagent (diphenylcarbazide and 1,10-phenanthroline) is studied. We have applied H-point standard addition method to determine concentrations of chromium (VI) and iron (III) from the mixture. The pure signals of complexes of chromium (VI) with diphenylcarbazide and iron (III) with the 1,10-phenanthroline and their calibration plots are previously carried out. We established the possibility of simultaneous determination of chromium (VI) and iron (III) in the different concentration ranges by H-point standard addition method. Correctness of determination of concentration by means of the offered technique is proved by "added-found" method for a series of mixtures with different ratios of concentration of chromium (VI) and iron (III). It is founded that the error of determination of concentration doesn't exceed 33%.

  17. Calibration of pulsed electroacoustic method considering electrode-dielectric interface status and porosity

    NASA Astrophysics Data System (ADS)

    Huang, Meng; Zhou, Yuanxiang; Chen, Weijiang; Lu, Licheng; Jin, Fubao; Huang, Jianwen

    2014-10-01

    Pulsed electroacoustic (PEA) method is widely used, but measurement conditions not always agree with the underlying PEA assumptions. This necessitates space charge recovery; however, existing research only addresses the attenuation and dispersion in lossy media. The effects of electrode-dielectric interface contact status and porosity on the accuracy of space charge distribution are discussed in the present article. It is shown that the presence of nonlinear interface and porosity can introduce severe error. However, because the properties of acoustic propagation of waves (which are generated from charge on the electrode and in the bulk) are different, the conventional recovery algorithm is no longer suitable for calibrating the charge density. To obtain accurate space charge profiles, it is necessary to eliminate these effects. A method has been proposed which is based on the original measurement process. The validity of the proposed method was tested by reasonable post-recovery electric field distributions.

  18. Method to calibrate phase fluctuation in polarization-sensitive swept-source optical coherence tomography

    NASA Astrophysics Data System (ADS)

    Lu, Zenghai; Kasaragod, Deepa K.; Matcher, Stephen J.

    2011-07-01

    We present a phase fluctuation calibration method for polarization-sensitive swept-source optical coherence tomography (PS-SS-OCT) using continuous polarization modulation. The method uses a low-voltage broadband polarization modulator driven by a synchronized sinusoidal burst waveform rather than an asynchronous waveform, together with the removal of the global phases of the measured Jones matrices by the use of matrix normalization. This makes it possible to average the measured Jones matrices to remove the artifact due to the speckle noise of the signal in the sample without introducing auxiliary optical components into the sample arm. This method was validated on measurements of an equine tendon sample by the PS-SS-OCT system.

  19. An improved method to accurately calibrate the gantry angle indicators of the radiotherapy linear accelerators

    NASA Astrophysics Data System (ADS)

    Chang, Liyun; Ho, Sheng-Yow; Du, Yi-Chun; Lin, Chih-Ming; Chen, Tainsong

    2007-06-01

    The calibration of the gantry angle indicator is an important and basic quality assurance (QA) item for the radiotherapy linear accelerator. In this study, we propose a new and practical method, which uses only the digital level, V-film, and general solid phantoms. By taking the star shot only, we can accurately calculate the true gantry angle according to the geometry of the film setup. The results on our machine showed that the gantry angle was shifted by -0.11° compared with the digital indicator, and the standard deviation was within 0.05°. This method can also be used for the simulator. In conclusion, this proposed method could be adopted as an annual QA item for mechanical QA of the accelerator.

  20. RAPID COMMUNICATION: Traceability of acoustic emission measurements using energy calibration methods

    NASA Astrophysics Data System (ADS)

    Yan, T.; Jones, B. E.

    2000-11-01

    Passive acoustic emission (AE) methods are becoming useful tools for integrity assessment of structures, monitoring of industrial processes and machines, and materials characterization. Unfortunately, there are no measurement standards for estimating the absolute strength of the AE sources. The lack of standardization makes it very difficult to compare the results obtained in different laboratories or on different structures, and to obtain meaningful repeatability of measurements. Therefore, current methods only give a qualitative rather than quantitative indication of the change of state of structure or process. This communication outlines a way of calibrating AE transducer systems in situ using a pulsed-laser-generated thermoelastic AE energy source or a bouncing-ball-generated elastic impact AE energy source. The methods presented here should enable traceable measurement standards to be established for AE.

  1. Coordinates calibration method in a robotic remanufacturing measurement system based on linear laser scanner

    NASA Astrophysics Data System (ADS)

    Shen, C. D.; Zhu, S.; Li, C.; Liang, Y. Y.

    2009-07-01

    In robotic remanufacturing measurement system, the 3D laser scanner is arranged by the robot and the object scanned is mounted on a turntable. This paper deals with the method of calibrating the relationship between the scanner coordinate and the robot Tool0 and furthermore locating the center axis of the turntable. The data of Tool0 can be directly obtained denoting its relationship with the robot base coordinate. So, the new methods of coordinate's transformation are effectively developed. Moreover some motivated experiments and optimized programs are designed for realizing process stabilization and reliability. This paper detailed explains the basic algorithm theory, practical operation instructions, the experiment data analysis, and etc. Theory deduction and experiments show the new methods are reasonable and efficient.

  2. Mass estimation of shaped charge jets from x-ray shadow graph with new calibration curve method

    NASA Astrophysics Data System (ADS)

    Saito, Fumikazu; Kishimura, Hiroaki; Kumakura, Akira; Sakai, Shun

    2015-06-01

    In order to assess the penetration capability of the Al and Cu metal jets against a bumper structure (such as Al plate and/or Al block), we measured the initial formation process of the metal jets generated from conical shaped charge device. The shaped charge device configurations employed in the experimental and numerical investigations have conical aluminum (and cupper) liner and steel casing with PBX explosive charge. The profile and velocity of the jets are measured with flash x-ray and x-ray film system. The mass of the jet tip are estimated from x-ray images by a calibration curve method proposed by our group. Al targets are used to evaluate a penetration performance of the jets. Additionally, we have simulated the initial formation process of the shaped charge jets with Autodyne-2D hydrodynamic code, which proposed important data to compare the experimental one.

  3. A standardized method for the calibration of thermodynamic data for the prediction of gas chromatographic retention times.

    PubMed

    McGinitie, Teague M; Ebrahimi-Najafabadi, Heshmatollah; Harynuk, James J

    2014-02-21

    A new method for calibrating thermodynamic data to be used in the prediction of analyte retention times is presented. The method allows thermodynamic data collected on one column to be used in making predictions across columns of the same stationary phase but with varying geometries. This calibration is essential as slight variances in the column inner diameter and stationary phase film thickness between columns or as a column ages will adversely affect the accuracy of predictions. The calibration technique uses a Grob standard mixture along with a Nelder-Mead simplex algorithm and a previously developed model of GC retention times based on a three-parameter thermodynamic model to estimate both inner diameter and stationary phase film thickness. The calibration method is highly successful with the predicted retention times for a set of alkanes, ketones and alcohols having an average error of 1.6s across three columns. PMID:24484693

  4. Novel method of calibration with restrictive constraints for stereo-vision system

    NASA Astrophysics Data System (ADS)

    Cui, Jiashan; Huo, Ju; Yang, Ming

    2016-05-01

    Regarding the calibration of a stereo vision measurement system, this paper puts forward a new bundle adjustment algorithm based on the stereo vision camera calibration method. Multiple-view geometric constraints and a bundle adjustment algorithm are used to optimize the inner and outer parameters of the camera accurately. A fixed relative constraint relationship between cameras is introduced. We have improved the normal equation construction process of the traditional bundle adjustment method, so that each iteration process occurs just outside the parameters of two images that are taken by a camera that has been optimized to better integrate two cameras bound together as one camera. The relationship between the fixed relative constraints can effectively increase the number of superfluous observations of the adjustment system and optimize higher accuracy while reducing the dimension of the normal matrix; it means that each iteration will reduce the time required. Simulation and actual experimental results show the superior performance of the proposed approach in terms of robustness and accuracy, and our approach also can be extended to stereo-vision system with more than two cameras.

  5. Generic precise augmented reality guiding system and its calibration method based on 3D virtual model.

    PubMed

    Liu, Miao; Yang, Shourui; Wang, Zhangying; Huang, Shujun; Liu, Yue; Niu, Zhenqi; Zhang, Xiaoxuan; Zhu, Jigui; Zhang, Zonghua

    2016-05-30

    Augmented reality system can be applied to provide precise guidance for various kinds of manual works. The adaptability and guiding accuracy of such systems are decided by the computational model and the corresponding calibration method. In this paper, a novel type of augmented reality guiding system and the corresponding designing scheme are proposed. Guided by external positioning equipment, the proposed system can achieve high relative indication accuracy in a large working space. Meanwhile, the proposed system is realized with a digital projector and the general back projection model is derived with geometry relationship between digitized 3D model and the projector in free space. The corresponding calibration method is also designed for the proposed system to obtain the parameters of projector. To validate the proposed back projection model, the coordinate data collected by a 3D positioning equipment is used to calculate and optimize the extrinsic parameters. The final projecting indication accuracy of the system is verified with subpixel pattern projecting technique. PMID:27410124

  6. Laser ablation methods for analysis of urinary calculi: Comparison study based on calibration pellets

    NASA Astrophysics Data System (ADS)

    Štěpánková, K.; Novotný, K.; Vašinová Galiová, M.; Kanický, V.; Kaiser, J.; Hahn, D. W.

    2013-03-01

    Methods based on laser ablation, such as Laser-Induced Breakdown Spectroscopy (LIBS) and Laser-Ablation Inductively Coupled Plasma Mass/Optical Emission Spectrometry (LA-ICP-MS/OES) are particularly suitable for urinary calculi bulk and micro analysis. Investigation of spatial distribution of matrix and trace elements can help to explain their emergence and growth. However, quantification is still very problematic and these methods are often used only for qualitative elemental mapping. There are no commercially available standards, which would correspond to the urinary calculi matrix. Internal standardization is also difficult, mainly due to different crystalline phases in one kidney stone. The aim of this study is to demonstrate the calibration capabilities and examine the limitations of laser ablation based techniques. Calibration pellets were prepared from powdered human urinary calculi with phosphate, oxalate and urate matrix. For this comparative study, the most frequently used laser-ablation based analytical techniques were chosen, such as LIBS and LA-ICP-MS. Moreover, some alternative techniques such as simultaneous LIBS-LA-ICP-OES and laser ablation LA-LIBS were also utilized.

  7. Methods for Calibrating Basin-Wide Hydroacoustic Propagation in the Indian Ocean

    SciTech Connect

    Blackman, D; de Groot-Hedlin, C; Orcutt, J A; Harben, P H; Clarke, D B; Ramirez, A L

    2004-10-11

    This collaborative project was designed to test and compare methods for achieving full ocean basin propagation of hydroacoustic signals in the 5-100 Hz frequency band. Plans for a systematic calibration of the International Monitoring System (IMS) for nuclear testing were under consideration in 2000/2001. The results from this project provide information to guide such planning for future ocean basin calibration work. Several acoustic source types were tested during two sea-going experiments and most were successful at generating signals that propagated hundreds to thousands of km to be recorded at the Indian Ocean IMS hydrophone stations. Development and numerical modeling of imploding glass sphere sources was one component of this testing. The intent was to design a relatively simple-to-use source that is not subject to restrictions that can limit use of explosive charges, but whose signal is large enough to propagate 100-1000's km range. Analysis of IMS hydrophone data recording during the experiments was used to illustrate the extent of energy loss during signal propagation and to assess the accuracy with which the small acoustic sources could be located using methods typically employed for nuclear monitoring.

  8. Primary reciprocity-based method for calibration of hydrophone magnitude and phase sensitivity: complete tests at frequencies from 1 to 7 MHz.

    PubMed

    Oliveira, E G; Costa-Felix, R P B; Machado, J C

    2015-04-01

    A primary reciprocity-based method for calibration of hydrophone magnitude and phase sensitivity is proposed. The method starts determining the transmit transfer function of an auxiliary transducer, based on the self-reciprocity method and using a stainless steel cylinder as reflecting target. Afterwards, the hydrophone, to be calibrated, is positioned facing the auxiliary transducer. The pressure field waveform, calculated at the hydrophone spot and based on the transmit transfer function of an auxiliary transducer, is used together with the output end of cable voltage waveform signal from the hydrophone to yield the calibrated hydrophone sensitivity. The method was tested with two similar membrane hydrophones, at frequencies within the 1.0-7.0 MHz range, in steps of 1.0 MHz. Results for magnitude sensitivity agree, within a confidence level of 95%, with those from previous calibration of same hydrophones at the National Physical Laboratory, in the UK (Enor⩽1.0). Phase sensitivity results agree with literature reported ones concerning the achieved uncertainty. Additionally, the phase sensitivities measured at 5.0 MHz for two similar hydrophones and employing two distinct auxiliary transducers presented no statistical significant difference. The method yielded a relative expanded uncertainty (p=0.95) for the sensitivity magnitude ranging between 6.6 and 7.0%, and an expanded uncertainty (p=0.95) ranging between 12° and 17° for the phase sensitivity. The results obtained so far lead to conclude that the proposed hydrophone calibration method is a validated alternative to the different existing methods. PMID:25578371

  9. Multilevel additive Schwarz method for the h-p version of the Galerkin boundary element method

    NASA Astrophysics Data System (ADS)

    Heuer, N.; Stephan, E. P.; Tran, T.

    1998-04-01

    We study a multilevel additive Schwarz method for the h-p version of the Galerkin boundary element method with geometrically graded meshes. Both hypersingular and weakly singular integral equations of the first kind are considered. As it is well known the h-p version with geometric meshes converges exponentially fast in the energy-norm. However, the condition number of the Galerkin matrix in this case blows up exponentially in the number of unknowns M. We prove that the condition number kappa(P) of the multilevel additive Schwarz operator behaves like O(root Mlog(2) M). Asa direct consequence of this we also give the results for the 2-level preconditioner and also for the h-p version with quasi-uniform meshes. Numerical results supporting our theory are presented.

  10. Online Calibration Methods for the DINA Model with Independent Attributes in CD-CAT

    ERIC Educational Resources Information Center

    Chen, Ping; Xin, Tao; Wang, Chun; Chang, Hua-Hua

    2012-01-01

    Item replenishing is essential for item bank maintenance in cognitive diagnostic computerized adaptive testing (CD-CAT). In regular CAT, online calibration is commonly used to calibrate the new items continuously. However, until now no reference has publicly become available about online calibration for CD-CAT. Thus, this study investigates the…

  11. A Measuring System for Well Logging Attitude and a Method of Sensor Calibration

    PubMed Central

    Ren, Yong; Wang, Yangdong; Wang, Mijian; Wu, Sheng; Wei, Biao

    2014-01-01

    This paper proposes an approach for measuring the azimuth angle and tilt angle of underground drilling tools with a MEMS three-axis accelerometer and a three-axis fluxgate sensor. A mathematical model of well logging attitude angle is deduced based on combining space coordinate transformations and algebraic equations. In addition, a system implementation plan of the inclinometer is given in this paper, which features low cost, small volume and integration. Aiming at the sensor and assembly errors, this paper analyses the sources of errors, and establishes two mathematical models of errors and calculates related parameters to achieve sensor calibration. The results show that this scheme can obtain a stable and high precision azimuth angle and tilt angle of drilling tools, with the deviation of the former less than ±1.4° and the deviation of the latter less than ±0.1°. PMID:24859028

  12. A measuring system for well logging attitude and a method of sensor calibration.

    PubMed

    Ren, Yong; Wang, Yangdong; Wang, Mijian; Wu, Sheng; Wei, Biao

    2014-01-01

    This paper proposes an approach for measuring the azimuth angle and tilt angle of underground drilling tools with a MEMS three-axis accelerometer and a three-axis fluxgate sensor. A mathematical model of well logging attitude angle is deduced based on combining space coordinate transformations and algebraic equations. In addition, a system implementation plan of the inclinometer is given in this paper, which features low cost, small volume and integration. Aiming at the sensor and assembly errors, this paper analyses the sources of errors, and establishes two mathematical models of errors and calculates related parameters to achieve sensor calibration. The results show that this scheme can obtain a stable and high precision azimuth angle and tilt angle of drilling tools, with the deviation of the former less than ±1.4° and the deviation of the latter less than ±0.1°. PMID:24859028

  13. New method for atmospheric calibration at the Pierre Auger Observatory using FRAM, a robotic astronomical telescope

    NASA Astrophysics Data System (ADS)

    Trávníček, Petr; Benzvi, Segev; Boháčová, Martina; Connolly, Brian; Grygar, Jiří; Hrabovský, Miroslav; Kárová, Tatiana; Mandát, Dušan; Nečesal, Petr; Nosek, Dalibor; Nožka, Libor; Palatka, Miroslav; sPech, Miroslav; Prouza, Michael; Řídký, Jan; Schovánek, Petr; Šmída, Radomír; Vitale, Primo; Westerhoff, Stefan

    FRAM - F/(Ph)otometric Robotic Atmospheric Monitor is the latest addition to the atmospheric monitoring instruments of the Pierre Auger Observatory. An optical telescope equipped with CCD camera and photometer, it automatically observes a set of selected standard stars and a calibrated terrestrial source. Primarily, the wavelength dependence of the attenuation is derived and the comparison between its vertical values (for stars) and horizontal values (for the terrestrial source) is made. Further, the integral vertical aerosol optical depth can be obtained. A secondary program of the instrument, the detection of optical counterparts of gamma-ray bursts, has already proven successful. The hardware setup, software system, data taking procedures, and first analysis results are described in this paper.

  14. Nebulizer calibration using lithium chloride: an accurate, reproducible and user-friendly method.

    PubMed

    Ward, R J; Reid, D W; Leonard, R F; Johns, D P; Walters, E H

    1998-04-01

    Conventional gravimetric (weight loss) calibration of jet nebulizers overestimates their aerosol output by up to 80% due to unaccounted evaporative loss. We examined two methods of measuring true aerosol output from jet nebulizers. A new adaptation of a widely available clinical assay for lithium (determined by flame photometry, LiCl method) was compared to an existing electrochemical method based on fluoride detection (NaF method). The agreement between the two methods and the repeatability of each method were examined. Ten Mefar jet nebulizers were studied using a Mefar MK3 inhalation dosimeter. There was no significant difference between the two methods (p=0.76) with mean aerosol output of the 10 nebulizers being 7.40 mg x s(-1) (SD 1.06; range 5.86-9.36 mg x s(-1)) for the NaF method and 7.27 mg x s(-1) (SD 0.82; range 5.52-8.26 mg x s(-1)) for the LiCl method. The LiCl method had a coefficient of repeatability of 13 mg x s(-1) compared with 3.7 mg x s(-1) for the NaF method. The LiCl method accurately measured true aerosol output and was considerably easier to use. It was also more repeatable, and hence more precise, than the NaF method. Because the LiCl method uses an assay that is routinely available from hospital biochemistry laboratories, it is easy to use and, thus, can readily be adopted by busy respiratory function departments. PMID:9623700

  15. A validated method for the quantitation of 1,1-difluoroethane using a gas in equilibrium method of calibration.

    PubMed

    Avella, Joseph; Lehrer, Michael; Zito, S William

    2008-10-01

    1,1-Difluoroethane (DFE), also known as Freon 152A, is a member of a class of compounds known as halogenated hydrocarbons. A number of these compounds have gained notoriety because of their ability to induce rapid onset of intoxication after inhalation exposure. Abuse of DFE has necessitated development of methods for its detection and quantitation in postmortem and human performance specimens. Furthermore, methodologies applicable to research studies are required as there have been limited toxicokinetic and toxicodynamic reports published on DFE. This paper describes a method for the quantitation of DFE using a gas chromatography-flame-ionization headspace technique that employs solventless standards for calibration. Two calibration curves using 0.5 mL whole blood calibrators which ranged from A: 0.225-1.350 to B: 9.0-180.0 mg/L were developed. These were evaluated for linearity (0.9992 and 0.9995), limit of detection of 0.018 mg/L, limit of quantitation of 0.099 mg/L (recovery 111.9%, CV 9.92%), and upper limit of linearity of 27,000.0 mg/L. Combined curve recovery results of a 98.0 mg/L DFE control that was prepared using an alternate technique was 102.2% with CV of 3.09%. No matrix interference was observed in DFE enriched blood, urine or brain specimens nor did analysis of variance detect any significant differences (alpha = 0.01) in the area under the curve of blood, urine or brain specimens at three identical DFE concentrations. The method is suitable for use in forensic laboratories because validation was performed on instrumentation routinely used in forensic labs and due to the ease with which the calibration range can be adjusted. Perhaps more importantly it is also useful for research oriented studies because the removal of solvent from standard preparation eliminates the possibility for solvent induced changes to the gas/liquid partitioning of DFE or chromatographic interference due to the presence of solvent in specimens. PMID:19007521

  16. Method for in-situ restoration of plantinum resistance thermometer calibration

    SciTech Connect

    Carroll, Radford M.

    1989-01-01

    A method is provided for in-situ restoration of platinum resistance thermometers (PRT's) that have undergone surface oxide contamination and/or strain-related damage causing decalibration. The method, which may be automated using a programmed computer control arrangement, consists of applying a dc heating current to the resistive sensing element of the PRT of sufficient magnitude to heat the element to an annealing temperature and maintaining the temperature for a specified period to restore the element to a stress-free calibration condition. The process anneals the sensing element of the PRT without subjecting the entire PRT assembly to the annealing temperature and may be used in the periodic maintenance of installed PRT's.

  17. Method for in-situ restoration of platinum resistance thermometer calibration

    DOEpatents

    Carroll, R.M.

    1987-10-23

    A method is provided for in-situ restoration of platinum resistance thermometers (PRT's) that have undergone surface oxide contamination and/or stain-related damage causing decalibration. The method, which may be automated using a programmed computer control arrangement, consists of applying a dc heating current to the resistive sensing element of the PRT of sufficient magnitude to heat the element to an annealing temperature and maintaining the temperature for a specified period to restore the element to a stress-free calibration condition. The process anneals the sensing element of the PRT without subjecting the entire PRT assembly to the annealing temperature and may be used in the periodic maintenance of installed PRT's. 1 fig.

  18. A Novel Error Model of Optical Systems and an On-Orbit Calibration Method for Star Sensors

    PubMed Central

    Wang, Shuang; Geng, Yunhai; Jin, Rongyu

    2015-01-01

    In order to improve the on-orbit measurement accuracy of star sensors, the effects of image-plane rotary error, image-plane tilt error and distortions of optical systems resulting from the on-orbit thermal environment were studied in this paper. Since these issues will affect the precision of star image point positions, in this paper, a novel measurement error model based on the traditional error model is explored. Due to the orthonormal characteristics of image-plane rotary-tilt errors and the strong nonlinearity among these error parameters, it is difficult to calibrate all the parameters simultaneously. To solve this difficulty, for the new error model, a modified two-step calibration method based on the Extended Kalman Filter (EKF) and Least Square Methods (LSM) is presented. The former one is used to calibrate the main point drift, focal length error and distortions of optical systems while the latter estimates the image-plane rotary-tilt errors. With this calibration method, the precision of star image point position influenced by the above errors is greatly improved from 15.42% to 1.389%. Finally, the simulation results demonstrate that the presented measurement error model for star sensors has higher precision. Moreover, the proposed two-step method can effectively calibrate model error parameters, and the calibration precision of on-orbit star sensors is also improved obviously. PMID:26703599

  19. A Novel Error Model of Optical Systems and an On-Orbit Calibration Method for Star Sensors.

    PubMed

    Wang, Shuang; Geng, Yunhai; Jin, Rongyu

    2015-01-01

    In order to improve the on-orbit measurement accuracy of star sensors, the effects of image-plane rotary error, image-plane tilt error and distortions of optical systems resulting from the on-orbit thermal environment were studied in this paper. Since these issues will affect the precision of star image point positions, in this paper, a novel measurement error model based on the traditional error model is explored. Due to the orthonormal characteristics of image-plane rotary-tilt errors and the strong nonlinearity among these error parameters, it is difficult to calibrate all the parameters simultaneously. To solve this difficulty, for the new error model, a modified two-step calibration method based on the Extended Kalman Filter (EKF) and Least Square Methods (LSM) is presented. The former one is used to calibrate the main point drift, focal length error and distortions of optical systems while the latter estimates the image-plane rotary-tilt errors. With this calibration method, the precision of star image point position influenced by the above errors is greatly improved from 15.42% to 1.389%. Finally, the simulation results demonstrate that the presented measurement error model for star sensors has higher precision. Moreover, the proposed two-step method can effectively calibrate model error parameters, and the calibration precision of on-orbit star sensors is also improved obviously. PMID:26703599

  20. A new method for the absolute radiance calibration for UV/vis measurements of scattered sun light

    NASA Astrophysics Data System (ADS)

    Wagner, T.; Beirle, S.; Dörner, S.; Penning de Vries, M.; Remmers, J.; Rozanov, A.; Shaiganfar, R.

    2015-05-01

    Absolute radiometric calibrations are important for measurements of the atmospheric spectral radiance. Such measurements can be used to determine actinic fluxes, the properties of aerosols and clouds and the short wave energy budget. Conventional calibration methods in the laboratory are based on calibrated light sources and reflectors and are expensive, time consuming and subject to relatively large uncertainties. Also, the calibrated instruments might change during transport from the laboratory to the measurement sites. Here we present a new calibration method for UV/vis instruments that measure the spectrally resolved sky radiance, like for example zenith sky Differential Optical Absorption Spectroscopy (DOAS-) instruments or Multi-AXis (MAX-) DOAS instruments. Our method is based on the comparison of the solar zenith angle dependence of the measured zenith sky radiance with radiative transfer simulations. For the application of our method clear sky measurements during periods with almost constant aerosol optical depth are needed. The radiative transfer simulations have to take polarisation into account. We show that the calibration results are almost independent from the knowledge of the aerosol optical properties and surface albedo, which causes a rather small uncertainty of about <7%. For wavelengths below about 330 nm it is essential that the ozone column density during the measurements is constant and known.

  1. An Automated Method of MFRSR Calibration for Aerosol Optical Depth Analysis with Application to an Asian Dust Outbreak Over the United States

    SciTech Connect

    Augustine, J. A.; Cornwall, C. R.; Hodges, G. B.; Long, Charles N.; Medina, C. I.; DeLuisi, J. J.

    2003-02-01

    Modern robotic spectral solar instruments designed for retrievals of aerosol optical depth (AOD), such as the Multi-Filter Rotating Shadowband Radiometer (MFRSR ) (Harrison et al. 1994), usually operate in an unattended mode. Thus their raw data sets sample a wide range of atmospheric conditions, most of which are undesirable for aerosol optical depth analysis. In addition, these instruments are often not calibrated for absolute irradiance, and must be calibrated for AOD analysis from their own operational data. For AOD retrievals, this involves extrapolation to the value that the instrument would measure before the sun's beam enters the earth's atmosphere, i.e., the extraterrestrial, or zero air mass signal (I 0). This value is inferred via the Langley method (Shaw 1983). Recently, a method that utilizes component solar measurements (direct and diffuse) to identify totally clear-sky and non-hazy periods (Long and Ackerman 2000) has been used successfully to screen MFRSR data for spectral solar measurements suitable for calibration Langley plots. This method was tested in a proof-of-concept mode on a two-month period during the Spring of 2001 with data from the Table Mountain SURF RAD station near Boulder, Colo. The resultant calibration is subsequently applied to an Asian dust event that occurred within that period, and verified with independent aerosol optical depth measurements from a nearby MFRSR and an automated sun photometer.

  2. Sensitivity and Calibration of Non-Destructive Evaluation Method That Uses Neural-Net Processing of Characteristic Fringe Patterns

    NASA Technical Reports Server (NTRS)

    Decker, Arthur J.; Weiland, Kenneth E.

    2003-01-01

    This paper answers some performance and calibration questions about a non-destructive-evaluation (NDE) procedure that uses artificial neural networks to detect structural damage or other changes from sub-sampled characteristic patterns. The method shows increasing sensitivity as the number of sub-samples increases from 108 to 6912. The sensitivity of this robust NDE method is not affected by noisy excitations of the first vibration mode. A calibration procedure is proposed and demonstrated where the output of a trained net can be correlated with the outputs of the point sensors used for vibration testing. The calibration procedure is based on controlled changes of fastener torques. A heterodyne interferometer is used as a displacement sensor for a demonstration of the challenges to be handled in using standard point sensors for calibration.

  3. Linear and Nonlinear Calibration Methods for Predicting Mechanical Properties of Polypropylene Pellets Using Raman Spectroscopy.

    PubMed

    Banquet-Terán, Julio; Johnson-Restrepo, Boris; Hernández-Morelo, Alveiro; Ropero, Jorge; Fontalvo-Gomez, Miriam; Romañach, Rodolfo J

    2016-07-01

    A nondestructive and faster methodology to quantify mechanical properties of polypropylene (PP) pellets, obtained from an industrial plant, was developed with Raman spectroscopy. Raman spectra data were obtained from several types of samples such as homopolymer PP, random ethylene-propylene copolymer, and impact ethylene-propylene copolymer. Multivariate calibration models were developed by relating the changes in the Raman spectra to mechanical properties determined by ASTM tests (Young's traction modulus, tensile strength at yield, elongation at yield on traction, and flexural modulus at 1% secant). Several strategies were evaluated to build robust models including the use of preprocessing methods (baseline correction, vector normalization, de-trending, and standard normal variate), selecting the best subset of wavelengths to model property response and discarding irrelevant variables by applying genetic algorithm (GA). Linear multivariable models were investigated such as partial least square regression (PLS) and PLS with genetic algorithm (GA-PLS) while nonlinear models were implemented with artificial neural network (ANN) preceded by GA (GA-ANN). The best multivariate calibration models were obtained when a combination of genetic algorithms and artificial neural network were used on Raman spectral data with relative standard errors (%RSE) from 0.17 to 0.41 for training and 0.42 to 0.88% validation data sets. PMID:27287847

  4. Method and apparatus usable with a calibration device for measuring the radioactivity of a sample

    SciTech Connect

    Suzuki, A.; Allardice, W.D. Jr.

    1985-03-19

    Method and apparatus of determining the radioactivity of a sample. The sample is contained in a vial located within a cannister having a body portion releasably engaged with a bottom portion. The apparatus includes a chuck for engaging and holding the bottom portion of the cannister and a lifting ring engageable with the body portion so that rotation of the lifting ring controls engagement and disengagement between the body and bottom portions of the cannister. After the chuck and cannister have been placed in a calibration device, the body portion is disengaged from the bottom portion by rotation of the lifting ring. The body portion is then moved by lifting the lifting ring so that the container is exposed and the radioactivity of the sample within the container can be measured. Subsequent lowering and rotation of the lifting ring reengages the body with the bottom of the cannister so that the sample is again shielded. The cannister and chuck are then removed from the calibration device.

  5. Ozone Correction for AM0 Calibrated Solar Cells for the Aircraft Method

    NASA Technical Reports Server (NTRS)

    Snyder, David B.; Scheiman, David A.; Jenkins, Phillip P.; Rieke, William J.; Blankenship, Kurt S.

    2002-01-01

    The aircraft solar cell calibration method has provided cells calibrated to space conditions for 37 years. However, it is susceptible to systematic errors due to ozone concentrations in the stratosphere. The present correction procedure applies a 1 percent increase to the measured I(sub SC) values. High band-gap cells are more sensitive to ozone absorbed wavelengths (0.4 to 0.8 microns) so it becomes important to reassess the correction technique. This paper evaluates the ozone correction to be 1+O3xFo, where O3 is the total ozone along the optical path, and Fo is 29.8 x 10(exp -6)/du for a Silicon solar cell, 42.6 x 10(exp -6)/du for a GaAs cell and 57.2 x 10(exp -6)/du for an InGaP cell. These correction factors work best to correct data points obtained during the flight rather than as a correction to the final result.

  6. A Single-Vector Force Calibration Method Featuring the Modern Design of Experiments

    NASA Technical Reports Server (NTRS)

    Parker, P. A.; Morton, M.; Draper, N.; Line, W.

    2001-01-01

    This paper proposes a new concept in force balance calibration. An overview of the state-of-the-art in force balance calibration is provided with emphasis on both the load application system and the experimental design philosophy. Limitations of current systems are detailed in the areas of data quality and productivity. A unique calibration loading system integrated with formal experimental design techniques has been developed and designated as the Single-Vector Balance Calibration System (SVS). This new concept addresses the limitations of current systems. The development of a quadratic and cubic calibration design is presented. Results from experimental testing are compared and contrasted with conventional calibration systems. Analyses of data are provided that demonstrate the feasibility of this concept and provide new insights into balance calibration.

  7. Preliminary evaluation of a microwave-assisted metal-labeling strategy for quantification of peptides via RPLC-ICP-MS and the method of standard additions.

    PubMed

    Christopher, Steven J; Kilpatrick, Eric L; Yu, Lee L; Davis, W Clay; Adair, Blakely M

    2012-01-15

    NIST has performed preliminary research on applying a calibration methodology based on the method of standard additions to the quantification of peptides via reverse-phase liquid chromatography coupled to inductively coupled plasma mass spectrometry (RPLC-ICP-MS). A microwave-assisted lanthanide labeling procedure was developed and applied to derivatize peptides using the macrocyclic bifunctional chemical chelator DOTA (1,4,7,10-tetraazacyclododecane-1,4,7,10-tetraacetic acid), which significantly improved the lanthanide labeling yield and reduced reaction times compared to benchtop labeling procedures. Biomolecular MS technologies of matrix-assisted laser desorption ionization (MALDI)-MS and electrospray ionization (ESI)-MS/MS were used in concert with ICP-MS to confirm the results of microwave labeling, sample cleanup and standard additions experiments for several test peptides. The calibration scheme is outlined in detail and contextualized against complementary high accuracy calibration strategies currently employed for ICP-MS detection of biomolecules. Standard additions experiments using native, non-isotopic peptide calibrants confirm the simplicity of the scheme and the potential of applying a blending (recombined sample and spike) procedure, facilitating calibration via co-elution of lanthanide labeled peptides. Ways to improve and fully leverage the analytical methodology are highlighted. PMID:22265570

  8. Sediment Core Extrusion Method at Millimeter Resolution Using a Calibrated, Threaded-rod.

    PubMed

    Schwing, Patrick T; Romero, Isabel C; Larson, Rebekka A; O'Malley, Bryan J; Fridrik, Erika E; Goddard, Ethan A; Brooks, Gregg R; Hastings, David W; Rosenheim, Brad E; Hollander, David J; Grant, Guy; Mulhollan, Jim

    2016-01-01

    Aquatic sediment core subsampling is commonly performed at cm or half-cm resolution. Depending on the sedimentation rate and depositional environment, this resolution provides records at the annual to decadal scale, at best. An extrusion method, using a calibrated, threaded-rod is presented here, which allows for millimeter-scale subsampling of aquatic sediment cores of varying diameters. Millimeter scale subsampling allows for sub-annual to monthly analysis of the sedimentary record, an order of magnitude higher than typical sampling schemes. The extruder consists of a 2 m aluminum frame and base, two core tube clamps, a threaded-rod, and a 1 m piston. The sediment core is placed above the piston and clamped to the frame. An acrylic sampling collar is affixed to the upper 5 cm of the core tube and provides a platform from which to extract sub-samples. The piston is rotated around the threaded-rod at calibrated intervals and gently pushes the sediment out the top of the core tube. The sediment is then isolated into the sampling collar and placed into an appropriate sampling vessel (e.g., jar or bag). This method also preserves the unconsolidated samples (i.e., high pore water content) at the surface, providing a consistent sampling volume. This mm scale extrusion method was applied to cores collected in the northern Gulf of Mexico following the Deepwater Horizon submarine oil release. Evidence suggests that it is necessary to sample at the mm scale to fully characterize events that occur on the monthly time-scale for continental slope sediments. PMID:27585268

  9. A New Method for Calibrating Perceptual Salience across Dimensions in Infants: The Case of Color vs. Luminance

    ERIC Educational Resources Information Center

    Kaldy, Zsuzsa; Blaser, Erik A.; Leslie, Alan M.

    2006-01-01

    We report a new method for calibrating differences in perceptual salience across feature dimensions, in infants. The problem of inter-dimensional salience arises in many areas of infant studies, but a general method for addressing the problem has not previously been described. Our method is based on a preferential looking paradigm, adapted to…

  10. Using sequential self-calibration method to identify conductivity distribution: Conditioning on tracer test data

    USGS Publications Warehouse

    Hu, B.X.; He, C.

    2008-01-01

    An iterative inverse method, the sequential self-calibration method, is developed for mapping spatial distribution of a hydraulic conductivity field by conditioning on nonreactive tracer breakthrough curves. A streamline-based, semi-analytical simulator is adopted to simulate solute transport in a heterogeneous aquifer. The simulation is used as the forward modeling step. In this study, the hydraulic conductivity is assumed to be a deterministic or random variable. Within the framework of the streamline-based simulator, the efficient semi-analytical method is used to calculate sensitivity coefficients of the solute concentration with respect to the hydraulic conductivity variation. The calculated sensitivities account for spatial correlations between the solute concentration and parameters. The performance of the inverse method is assessed by two synthetic tracer tests conducted in an aquifer with a distinct spatial pattern of heterogeneity. The study results indicate that the developed iterative inverse method is able to identify and reproduce the large-scale heterogeneity pattern of the aquifer given appropriate observation wells in these synthetic cases. ?? International Association for Mathematical Geology 2008.

  11. Development of a New Calibration Method for an Ambient Ion Monitor Ion Chromatograph (AIM-IC)

    NASA Astrophysics Data System (ADS)

    Markovic, M.; Vandenboer, T.; Murphy, J. G.

    2009-05-01

    Fine atmospheric aerosols play an important role in the atmosphere as they alter the radiative balance of the Earth through direct and indirect climate effects, reduce visibility, participate in acid rain formation and affect human health. The motivation for chemically and temporally resolved measurements of fine aerosol composition has lead to the development of the Ambient Ion Monitor Ion Chromatograph (AIM-IC) system by Dionex/URG. This instrument is capable of simultaneously monitoring fine aerosols (<2.5μm) and associated precursor gases on a nearly continuous basis with a time resolution of 1 hour. The instrument utilizes a parallel-plate wet denuder with a constantly regenerated surface for collection of gases and a particle condensation chamber for the collection of aerosols. AIM-IC is capable of monitoring HCl(g), HONO(g), HNO3(g), SO2(g), NH3(g), Cl-, NO2-, NO3-, SO42-, NH4+ , and some water soluble organic acids and amines. Standard calibration of the AIM-IC is carried out by injecting a series of mixed standards directly onto the ion chromatographs, bypassing the sampling component of the instrument. This results in calculated detection limits on the order of 10-200 pptv for gases and 10-500 of ng/m3 for individual particle constituents when collecting at 3 L/min for 55 minutes. In this work, we present a new method for the calibration of the AIM-IC for both gas and particle collection that enables us to evaluate the entire system from size-selection to detection. This external calibration method is assessed for the gases HNO3(g), SO2(g), and NH3(g), and for particles containing (NH4)2SO4, NH4NO3, and Na2SO4. Quantitative collection of SO2 is found to require careful optimization of the H2O2 concentration of the denuder liquid, while the replacement of a cyclone with an impactor improves the sampling efficiency of NH3 and HNO3.

  12. PLEIADES-HR 1A&1B image quality commissioning: innovative geometric calibration methods and results

    NASA Astrophysics Data System (ADS)

    Greslou, Daniel; de Lussy, Françoise; Amberg, Virginie; Dechoz, Cécile; Lenoir, Florie; Delvit, Jean-Marc; Lebègue, Laurent

    2013-09-01

    PLEIADES earth observing system consists of two satellites designed to provide optical 70cm resolution images to civilian and defense users. The first Pleiades satellite 1A was launched on December 2011 while the second satellite Pleiades 1B was placed on orbit, one year after, on December 2012. The calibration operations and the assessment of the image of the two satellites have been performed by CNES Image Quality team during the called commissioning phase which took place after each launch and lasted each time less than 6 months. The geometric commissioning activities consist in assessing and improving the geometric quality of the images in order to meet very demanding requirements. This paper deals with the means used and methods applied, mainly the innovative ones, in order to manage these activities. It describes both their accuracy and their operational interest. Finally it gives the main results for geometric image quality performances of the PHR system.

  13. Methods for Optical Calibration of the BigBite Hadron Spectrometer

    SciTech Connect

    M. Mihovilovic, K. Allada, B.D. Anderson, J.R.M. Annand, T. Averett, A. Camsonne, R.W. Chan, J.-P. Chen, K. Chirapatpimol, C.W. de Jager, S. Gilad, D.J. Hamilton, J.-O. Hansen, D.W. Higinbotham, J. Juang, X. Jiang, G. Jin, W. Korsch, J.J. LeRose, R.A. Lindgren, N. Liyanage, E. Long, R. Michaels, B. Moffit, P. Monaghan, V. Nelyubin, B.E. Norum, E. Piasetzky, X. Qian, Y. Qiang, S. Riordan, G. Ron, G. Rosner, B. Sawatzky, M. Shabestari, A. Shahinyan, R. Shneor, S. Sirca, R. Subedi, V. Sulkosky, J.W. Watson, B. Wojtsekhowski, Y.-W. Zhang

    2012-09-01

    The techniques for optical calibration of Jefferson Lab's large-acceptance magnetic hadron spectrometer, BigBite, have been examined. The most consistent and stable results were obtained by using a method based on singular value decomposition. In spite of the complexity of the optics, the particles positions and momenta at the target have been precisely reconstructed from the coordinates measured in the detectors by means of a single back-tracing matrix. The technique is applicable to any similar magnetic spectrometer and any particle type. For 0.55 GeV/c protons, we have established the vertex resolution of 1.2 cm, angular resolutions of 7 mrad and 16 mrad (in-plane and out-of-plane, respectively), and a relative momentum resolution of 1.6%.

  14. PRESAGE® as a new calibration method for high intensity focused ultrasound therapy

    NASA Astrophysics Data System (ADS)

    Costa, M.; McErlean, C.; Rivens, I.; Adamovics, J.; Leach, M. O.; ter Haar, G.; Doran, S. J.

    2015-01-01

    High Intensity Focused ultrasound (HIFU) is a non-invasive cancer therapy that makes use of the mainly thermal effects of ultrasound to destroy tissue. In order to achieve reliable treatment planning, it is necessary to characterise the ultrasound source (transducer) and to understand how the wave propagates in tissue and the energy deposition in the focal region. This novel exploratory study investigated how HIFU affects PRESAGE®, an optical phantom used for radiotherapy dosimetry, which is potentially a rapid method of calibrating the transducer. Samples, of two different formulations, were exposed to focused ultrasound and imaged using Optical Computed Tomography. First results showed that, PRESAGE® changes colour on ultrasound exposure (darker green regions were observed) with the alterations being related to the acoustic power and sample composition. Future work will involve quantification of these alterations and understanding how to relate them to the mechanisms of action of HIFU.

  15. Methods for optical calibration of the BigBite hadron spectrometer

    NASA Astrophysics Data System (ADS)

    Mihovilovič, M.; Širca, S.; Allada, K.; Anderson, B. D.; Annand, J. R. M.; Averett, T.; Camsonne, A.; Chan, R. W.; Chen, J.-P.; Chirapatpimol, K.; de Jager, C. W.; Gilad, S.; Hamilton, D. J.; Hansen, J.-O.; Higinbotham, D. W.; Huang, J.; Jiang, X.; Jin, G.; Korsch, W.; LeRose, J. J.; Lindgren, R. A.; Liyanage, N.; Long, E.; Michaels, R.; Moffit, B.; Monaghan, P.; Nelyubin, V.; Norum, B. E.; Piasetzky, E.; Qian, X.; Qiang, Y.; Riordan, S.; Ron, G.; Rosner, G.; Sawatzky, B.; Shabestari, M.; Shahinyan, A.; Shneor, R.; Subedi, R.; Sulkosky, V.; Watson, J. W.; Zhang, Y.-W.

    2012-09-01

    The techniques for the optical calibration of Jefferson Lab's large-acceptance magnetic hadron spectrometer, BigBite, have been examined. The most consistent and stable results were obtained using a method based on singular value decomposition. In spite of the complexity of the optics, the particles' positions and momenta at the target have been precisely reconstructed from the coordinates measured in the detectors by means of a single back-tracing matrix. The technique is applicable to any similar magnetic spectrometer and any particle type. For 0.55 GeV/c protons, we have established a vertex resolution of 1.2 cm, angular resolutions of 7 mrad and 13 mrad (in-plane and out-of-plane, respectively), and a relative momentum resolution of 1.6%.

  16. Uncertainty on differential measurements and its reduction using the calibration by comparison method

    NASA Astrophysics Data System (ADS)

    Ospina, José; Canuto, Enrico

    2008-08-01

    The paper deals with the uncertainty of differential measurements, obtained from the subtraction of a pair of absolute measurements. It is shown that if the same sensor is used to perform both measurements, a model of the sensor will reveal a correlation component between the uncertainty of each absolute measurement, reducing the uncertainty on its subtraction. The procedure followed is based on the Gauss-Markov estimation method, showing that differential measurement uncertainty vanishes when the gradient to be measured is zero. If the two absolute measurements are to be performed using different sensors, a calibration by comparison between them will result in a similar uncertainty reduction. Finally, a simulated example based on commercially available thermistor data is included.

  17. Investigating temporal field sampling strategies for site-specific calibration of three soil moisture-neutron intensity parameterisation methods

    NASA Astrophysics Data System (ADS)

    Iwema, J.; Rosolem, R.; Baatz, R.; Wagener, T.; Bogena, H. R.

    2015-07-01

    The Cosmic-Ray Neutron Sensor (CRNS) can provide soil moisture information at scales relevant to hydrometeorological modelling applications. Site-specific calibration is needed to translate CRNS neutron intensities into sensor footprint average soil moisture contents. We investigated temporal sampling strategies for calibration of three CRNS parameterisations (modified N0, HMF, and COSMIC) by assessing the effects of the number of sampling days and soil wetness conditions on the performance of the calibration results while investigating actual neutron intensity measurements, for three sites with distinct climate and land use: a semi-arid site, a temperate grassland, and a temperate forest. When calibrated with 1 year of data, both COSMIC and the modified N0 method performed better than HMF. The performance of COSMIC was remarkably good at the semi-arid site in the USA, while the N0mod performed best at the two temperate sites in Germany. The successful performance of COSMIC at all three sites can be attributed to the benefits of explicitly resolving individual soil layers (which is not accounted for in the other two parameterisations). To better calibrate these parameterisations, we recommend in situ soil sampled to be collected on more than a single day. However, little improvement is observed for sampling on more than 6 days. At the semi-arid site, the N0mod method was calibrated better under site-specific average wetness conditions, whereas HMF and COSMIC were calibrated better under drier conditions. Average soil wetness condition gave better calibration results at the two humid sites. The calibration results for the HMF method were better when calibrated with combinations of days with similar soil wetness conditions, opposed to N0mod and COSMIC, which profited from using days with distinct wetness conditions. Errors in actual neutron intensities were translated to average errors specifically to each site. At the semi-arid site, these errors were below the

  18. Calibration of the RPC charge readout in the ARGO-YBJ experiment with the iso-gradient method

    NASA Astrophysics Data System (ADS)

    Bartoli, B.; Bernardini, P.; Bi, X. J.; Branchini, P.; Budano, A.; Camarri, P.; Cao, Z.; Cardarelli, R.; Catalanotti, S.; Chen, S. Z.; Chen, T. L.; Creti, P.; Cui, S. W.; Dai, B. Z.; D`Amone, A.; Danzengluobu; De Mitri, I.; D`Ettorre Piazzoli, B.; Di Girolamo, T.; Di Sciascio, G.; Feng, C. F.; Feng, Zhaoyang; Feng, Zhenyong; Gou, Q. B.; Guo, Y. Q.; He, H. H.; Hu, Haibing; Hu, Hongbo; Iacovacci, M.; Iuppa, R.; Jia, H. Y.; Labaciren; Li, H. J.; Liguori, G.; Liu, C.; Liu, J.; Liu, M. Y.; Lu, H.; Ma, L. L.; Ma, X. H.; Mancarella, G.; Mari, S. M.; Marsella, G.; Martello, D.; Mastroianni, S.; Montini, P.; Ning, C. C.; Panareo, M.; Perrone, L.; Pistilli, P.; Ruggieri, F.; Salvini, P.; Santonico, R.; Shen, P. R.; Sheng, X. D.; Shi, F.; Surdo, A.; Tan, Y. H.; Vallania, P.; Vernetto, S.; Vigorito, C.; Wang, H.; Wu, C. Y.; Wu, H. R.; Xue, L.; Yang, Q. Y.; Yang, X. C.; Yao, Z. G.; Yuan, A. F.; Zha, M.; Zhang, H. M.; Zhang, L.; Zhang, X. Y.; Zhang, Y.; Zhao, J.; Zhaxiciren; Zhaxisangzhu; Zhou, X. X.; Zhu, F. R.; Zhu, Q. Q.; Zizzi, G.

    2015-05-01

    The ARGO-YBJ experiment is a full coverage array of Resistive Plate Chambers (RPCs) with an active area of 5800 m2. In order to eliminate the response difference of the charge readout from the RPCs, a calibration procedure is carried out with the iso-gradient method. This method also allows the extension of the absolute calibration with the muon telescope including scintillation detectors to all the RPCs in the array. The overall systematic uncertainty in measurements of the number of particles by the RPCs is 10.7%. In general, the method gives results consistent with those from a totally different approach also used in the experiment.

  19. Calibration-based NUC Method in Real-time Based on IRFPA

    NASA Astrophysics Data System (ADS)

    Sheng, Meng; Xie, Juntang; Fu, Ziyuan

    The non-uniformity of Infrared Focal Plane Array (IRFPA) resulted from the limits of the detector's materials and the manufacturing process affects the performance of the staring IR imaging system. To address this problem, non-uniformity correction (NUC), applied for real-time resolution, is the important issue in the IR imaging information processing system. This thesis introduces method of non-uniformity correction. Considering the nonlinear character of IRFPA, the calibration-based polynomial NUC method is proposed in the hardware system. Comparing with the conventional NUC schemes, polynomial method can achieve better NUC performance and implement in real-time. The algorithm is designed based on System architecture for FPGA hardware, for which is the Xilinx ML402 platform dedicated for video processing, which consists of A/D and D/A converter, and Virtex-4 FPGA on the mother board. The polynomial method reduces the non-uniformity in the infrared image largely, implemented at real-time, as well as the advantage of wide dynamic range.

  20. A protocol and calibration method for accurate multi-camera field videography.

    PubMed

    Theriault, Diane H; Fuller, Nathan W; Jackson, Brandon E; Bluhm, Evan; Evangelista, Dennis; Wu, Zheng; Betke, Margrit; Hedrick, Tyson L

    2014-06-01

    Stereo videography is a powerful technique for quantifying the kinematics and behavior of animals, but it can be challenging to use in an outdoor field setting. We here present a workflow and associated software for performing calibration of cameras placed in a field setting and estimating the accuracy of the resulting stereoscopic reconstructions. We demonstrate the workflow through example stereoscopic reconstructions of bat and bird flight. We provide software tools for planning experiments and processing the resulting calibrations that other researchers may use to calibrate their own cameras. Our field protocol can be deployed in a single afternoon, requiring only short video clips of light, portable calibration objects. PMID:24577444

  1. The Development of Students' Use of Additive and Proportional Methods along Primary and Secondary School

    ERIC Educational Resources Information Center

    Fernandez, Ceneida; Llinares, Salvador; Van Dooren, Wim; De Bock, Dirk; Verschaffel, Lieven

    2012-01-01

    This study investigates the development of proportional and additive methods along primary and secondary school. In particular, it simultaneously investigates the use of additive methods in proportional word problems and the use of proportional methods in additive word problems. We have also studied the role played by integer and non-integer…

  2. An innovative method for coordinate measuring machine one-dimensional self-calibration with simplified experimental process

    NASA Astrophysics Data System (ADS)

    Fang, Cheng; Butler, David Lee

    2013-05-01

    In this paper, an innovative method for CMM (Coordinate Measuring Machine) self-calibration is proposed. In contrast to conventional CMM calibration that relies heavily on a high precision reference standard such as a laser interferometer, the proposed calibration method is based on a low-cost artefact which is fabricated with commercially available precision ball bearings. By optimizing the mathematical model and rearranging the data sampling positions, the experimental process and data analysis can be simplified. In mathematical expression, the samples can be minimized by eliminating the redundant equations among those configured by the experimental data array. The section lengths of the artefact are measured at arranged positions, with which an equation set can be configured to determine the measurement errors at the corresponding positions. With the proposed method, the equation set is short of one equation, which can be supplemented by either measuring the total length of the artefact with a higher-precision CMM or calibrating the single point error at the extreme position with a laser interferometer. In this paper, the latter is selected. With spline interpolation, the error compensation curve can be determined. To verify the proposed method, a simple calibration system was set up on a commercial CMM. Experimental results showed that with the error compensation curve uncertainty of the measurement can be reduced to 50%.

  3. A Comparative Study of Online Pretest Item Calibration/Scaling Methods in CAT. ACT Research Report Series.

    ERIC Educational Resources Information Center

    Ban, Jae-Chun; Hanson, Bradley A.; Wang, Tianyou; Yi, Qing; Harris, Deborah J.

    The purpose of this study was to compare and evaluate five online pretest item calibration/scaling methods in computerized adaptive testing (CAT): (1) the marginal maximum likelihood estimate with one-EM cycle (OEM); (2) the marginal maximum likelihood estimate with multiple EM cycles (MEM); (3) Stocking's Method A (M. Stocking, 1988); (4)…

  4. Calibration of Hinode/XRT for Coalignment II: Comparison of Three Different Methods

    NASA Astrophysics Data System (ADS)

    Yoshimura, Keiji; McKenzie, D. E.

    2013-07-01

    The coalignment of solar images from various observational instruments is an important step to study solar features/phenomena when using data acquired from multiple telescopes or wavelengths. We have been working on the calibration of XRT/Hinode to get coalignment parameters, i.e., plate scale, roll angle in the plane of the sky, and pointing information. We are developing a table of the parameters for every single XRT image, so that all users of XRT data can easily utilize our results for their coalignment. There are three different approaches to get the XRT pointing information: (1) applying the cross correlation technique to the X-ray images from XRT and the EUV images from AIA/SDO (2) combination of solar limb fitting and cross correlation between XRT full and partial disk images (3) using outputs from the sun sensor (Ultra Fine Sun Sensor, UFSS) on-board Hinode We discuss the advantages and disadvantages of each method and estimate the errors through the cross comparison. In preparing our table of coalignment coefficients we combine the strengths of each of these methods to provide the most reliable resource. We also discuss the accuracy of the plate scale of X-ray data from XRT, which can be derived from the comparison between XRT and AIA images.Abstract (2,250 Maximum Characters): The coalignment of solar images from various observational instruments is an important step to study solar features/phenomena when using data acquired from multiple telescopes or wavelengths. We have been working on the calibration of XRT/Hinode to get coalignment parameters, i.e., plate scale, roll angle in the plane of the sky, and pointing information. We are developing a table of the parameters for every single XRT image, so that all users of XRT data can easily utilize our results for their coalignment. There are three different approaches to get the XRT pointing information: (1) applying the cross correlation technique to the X-ray images from XRT and the EUV images from AIA

  5. A novel INS and Doppler sensors calibration method for long range underwater vehicle navigation.

    PubMed

    Tang, Kanghua; Wang, Jinling; Li, Wanli; Wu, Wenqi

    2013-01-01

    Since the drifts of Inertial Navigation System (INS) solutions are inevitable and also grow over time, a Doppler Velocity Log (DVL) is used to aid the INS to restrain its error growth. Therefore, INS/DVL integration is a common approach for Autonomous Underwater Vehicle (AUV) navigation. The parameters including the scale factor of DVL and misalignments between INS and DVL are key factors which limit the accuracy of the INS/DVL integration. In this paper, a novel parameter calibration method is proposed. An iterative implementation of the method is designed to reduce the error caused by INS initial alignment. Furthermore, a simplified INS/DVL integration scheme is employed. The proposed method is evaluated with both river trial and sea trial data sets. Using 0.03°/h(1σ) ring laser gyroscopes, 5 × 10-5 g(1σ) quartz accelerometers and DVL with accuracy 0.5% V ± 0.5 cm/s, INS/DVL integrated navigation can reach an accuracy of about 1‰ of distance travelled (CEP) in a river trial and 2‰ of distance travelled (CEP) in a sea trial. PMID:24169542

  6. A Novel INS and Doppler Sensors Calibration Method for Long Range Underwater Vehicle Navigation

    PubMed Central

    Tang, Kanghua; Wang, Jinling; Li, Wanli; Wu, Wenqi

    2013-01-01

    Since the drifts of Inertial Navigation System (INS) solutions are inevitable and also grow over time, a Doppler Velocity Log (DVL) is used to aid the INS to restrain its error growth. Therefore, INS/DVL integration is a common approach for Autonomous Underwater Vehicle (AUV) navigation. The parameters including the scale factor of DVL and misalignments between INS and DVL are key factors which limit the accuracy of the INS/DVL integration. In this paper, a novel parameter calibration method is proposed. An iterative implementation of the method is designed to reduce the error caused by INS initial alignment. Furthermore, a simplified INS/DVL integration scheme is employed. The proposed method is evaluated with both river trial and sea trial data sets. Using 0.03°/h(1σ) ring laser gyroscopes, 5 × 10−5 g(1σ) quartz accelerometers and DVL with accuracy 0.5% V ± 0.5 cm/s, INS/DVL integrated navigation can reach an accuracy of about 1‰ of distance travelled (CEP) in a river trial and 2‰ of distance travelled (CEP) in a sea trial. PMID:24169542

  7. Calibration of scattering and absorption properties of a liquid diffusive medium at NIR wavelengths. CW method.

    PubMed

    Martelli, Fabrizio; Zaccanti, Giovanni

    2007-01-22

    In spite of many progresses achieved both with theories and with experiments in studying light propagation through diffusive media, a reliable method for accurate measurements of the optical properties of diffusive media at NIR wavelengths is, in our opinion, still missing. It is therefore difficult to create a diffusive medium with well known optical properties to be used as a reference. In this paper we describe a method to calibrate the reduced scattering coefficient, mu'(s) , of a liquid diffusive medium and the absorption coefficient, mu(a), of an absorbing medium with a standard error smaller than 2% both on mu'(s) and on mu(a). The method is based on multidistance measurements of fluence into an infinite medium illuminated by a CW source. The optical properties are retrieved with simple inversion procedures (linear fits) exploiting the knowledge of the absorption coefficient of the liquid into which the diffuser and the absorber are dispersed. In this study Intralipid diluted in water has been used as diffusive medium and Indian ink as absorber. For a full characterization of these media measurements of collimated transmittance have also been carried out, from which the asymmetry factor of the scattering function of Intralipid and the single scattering albedo of Indian ink have been determined. PMID:19532267

  8. A new method for calibrating a boron isotope paleo-pH proxy using massive Porites corals

    NASA Astrophysics Data System (ADS)

    Kubota, Kaoru; Yokoyama, Yusuke; Ishikawa, Tsuyoshi; Suzuki, Atsushi

    2015-09-01

    The boron isotope ratio (δ11B) of marine biogenic carbonates can reconstruct pH and pCO2 of seawater, and potentially CO2 concentration in the atmosphere. To date, δ11B-pHSW calibration has been proposed via culturing experiments, where calcifying organisms are cultured under artificially acidified seawater. However, in scleractinian corals, reconstructed pH values using culture-based calibrations do not agree well with actual observations of seawater CO2 chemistry. Thus, another approach is needed to establish a more reliable calibration method. In this study, we established field-based calibrations for Chichijima and Tahiti, both located in subtropical gyres where surface seawater is close to CO2 equilibrium. We suggest a new approach to calibration of δ11B-pH in which the long-term δ11B variation of massive Porites corals is compared with the decreasing pH trend (i.e., ocean acidification) that has occurred since the Industrial Revolution. This calibration will offer a new avenue for studying seawater CO2 chemistry using coral δ11B in diverse settings, such as upwelling regions, coral reefs, and coastal areas.

  9. Array error calibration methods in downward-looking linear-array three-dimensional synthetic aperture radar

    NASA Astrophysics Data System (ADS)

    Tan, Weixian; Huang, Pingping; Han, Kuoye; Liu, Qi; Peng, Xueming

    2016-04-01

    In order to achieve high-precision three-dimensional (3-D) imaging with an airborne downward-looking linear-array 3-D synthetic aperture radar (LA-3D-SAR), a uniform virtual antenna array can be obtained by aperture synthesis of the cross-track sparse multiple-input-multiple-output array. However, the actual 3-D imaging quality is unavoidably degraded by array errors such as the multichannel amplitude-phase errors due to the nonideal antenna characteristics, and the virtual element position errors due to vibrations and motion measurement deviations. We investigate the effects of these errors on the forms and the degrees of image quality degradation and consider the use of corresponding calibration methods to eliminate the effects of errors. For the multichannel amplitude-phase errors, the target response is subject to an integrated sidelobe level increase introduced by the phase error, which can be calibrated based on external (parallel or point target) calibrators, as proposed in the paper. For the virtual element position errors, they mainly the result of contrast degradation and noise in the image along the cross-track direction and have little impact on the range and along-track directions. The imaging performance is more sensitive to the error component in the height direction as compared to other components, the precision requirement of which should be established as the calibration reference. A calibration method based on time-divided active calibrators is proposed to estimate and correct the virtual element position errors. Both numerical simulations and real data experiments have shown the validity of the analyses as well as the effectiveness of the proposed calibration methods.

  10. A Comparison of the Rasch Separate Calibration and Between-Fit Methods of Detecting Item Bias.

    ERIC Educational Resources Information Center

    Smith, Richard M.

    1996-01-01

    The separate calibration t-test approach of B. Wright and M. Stone (1979) and the common calibration between-fit approach of B. Wright, R. Mead, and R. Draba (1976) appeared to have similar Type I error rates and similar power to detect item bias within a Rasch framework. (SLD)

  11. Self-Calibrating and Remote Programmable Signal Conditioning Amplifier System and Method

    NASA Technical Reports Server (NTRS)

    Medelius, Pedro J. (Inventor); Hallberg, Carl G. (Inventor); Simpson, Howard J., III (Inventor); Thayer, Stephen W. (Inventor)

    1998-01-01

    A self-calibrating, remote programmable signal conditioning amplifier system employs information read from a memory attached to a measurement transducer for automatic calibration. The signal conditioning amplifier is self-calibrated on a continuous basis through use of a dual input path arrangement, with each path containing a multiplexer and a programmable amplifier. A digital signal processor controls operation of the system such that a transducer signal is applied to one of the input paths, while one or more calibration signals are applied to the second input path. Once the second path is calibrated, the digital signal processor switches the transducer signal to the second path. and then calibrates the first path. This process is continually repeated so that each path is calibrated on an essentially continuous basis. Dual output paths are also employed which are calibrated in the same manner. The digital signal processor also allows the implementation of a variety of digital filters which are either programmed into the system or downloaded by an operator, and performs up to eighth order linearization.

  12. Multi-Axis Accelerometer Calibration System

    NASA Technical Reports Server (NTRS)

    Finley, Tom; Parker, Peter

    2010-01-01

    A low-cost, portable, and simplified system has been developed that is suitable for in-situ calibration and/or evaluation of multi-axis inertial measurement instruments. This system overcomes facility restrictions and maintains or improves the calibration quality for users of accelerometer-based instruments with applications in avionics, experimental wind tunnel research, and force balance calibration applications. The apparatus quickly and easily positions a multi-axis accelerometer system into a precisely known orientation suitable for in-situ quality checks and calibration. In addition, the system incorporates powerful and sophisticated statistical methods, known as response surface methodology and statistical quality control. These methods improve calibration quality, reduce calibration time, and allow for increased calibration frequency, which enables the monitoring of instrument stability over time.

  13. Calibration of measurement sensitivities of multiple micro-cantilever dynamic modes in atomic force microscopy using a contact detection method

    SciTech Connect

    Liu Zhen; Jeong, Younkoo; Menq, Chia-Hsiang

    2013-02-15

    An accurate experimental method is proposed for on-spot calibration of the measurement sensitivities of multiple micro-cantilever dynamic modes in atomic force microscopy. One of the key techniques devised for this method is a reliable contact detection mechanism that detects the tip-surface contact instantly. At the contact instant, the oscillation amplitude of the tip deflection, converted to that of the deflection signal in laser reading through the measurement sensitivity, exactly equals to the distance between the sample surface and the cantilever base position. Therefore, the proposed method utilizes the recorded oscillation amplitude of the deflection signal and the base position of the cantilever at the contact instant for the measurement sensitivity calibration. Experimental apparatus along with various signal processing and control modules was realized to enable automatic and rapid acquisition of multiple sets of data, with which the calibration of a single dynamic mode could be completed in less than 1 s to suppress the effect of thermal drift and measurement noise. Calibration of the measurement sensitivities of the first and second dynamic modes of three micro-cantilevers having distinct geometries was successfully demonstrated. The dependence of the measurement sensitivity on laser spot location was also experimentally investigated. Finally, an experiment was performed to validate the calibrated measurement sensitivity of the second dynamic mode of a micro-cantilever.

  14. Calibration of measurement sensitivities of multiple micro-cantilever dynamic modes in atomic force microscopy using a contact detection method

    NASA Astrophysics Data System (ADS)

    Liu, Zhen; Jeong, Younkoo; Menq, Chia-Hsiang

    2013-02-01

    An accurate experimental method is proposed for on-spot calibration of the measurement sensitivities of multiple micro-cantilever dynamic modes in atomic force microscopy. One of the key techniques devised for this method is a reliable contact detection mechanism that detects the tip-surface contact instantly. At the contact instant, the oscillation amplitude of the tip deflection, converted to that of the deflection signal in laser reading through the measurement sensitivity, exactly equals to the distance between the sample surface and the cantilever base position. Therefore, the proposed method utilizes the recorded oscillation amplitude of the deflection signal and the base position of the cantilever at the contact instant for the measurement sensitivity calibration. Experimental apparatus along with various signal processing and control modules was realized to enable automatic and rapid acquisition of multiple sets of data, with which the calibration of a single dynamic mode could be completed in less than 1 s to suppress the effect of thermal drift and measurement noise. Calibration of the measurement sensitivities of the first and second dynamic modes of three micro-cantilevers having distinct geometries was successfully demonstrated. The dependence of the measurement sensitivity on laser spot location was also experimentally investigated. Finally, an experiment was performed to validate the calibrated measurement sensitivity of the second dynamic mode of a micro-cantilever.

  15. Technique for Radiometer and Antenna Array Calibration with Two Antenna Noise Diodes

    NASA Technical Reports Server (NTRS)

    Srinivasan, Karthik; Limaye, Ashutosh; Laymon, Charles; Meyer, Paul

    2011-01-01

    This paper presents a new technique to calibrate a microwave radiometer and phased array antenna system. This calibration technique uses a radiated noise source in addition to an injected noise sources for calibration. The plane of reference for this calibration technique is the face of the antenna and therefore can effectively calibration the gain fluctuations in the active phased array antennas. This paper gives the mathematical formulation for the technique and discusses the improvements brought by the method over the existing calibration techniques.

  16. Towards monolithic scintillator based TOF-PET systems: practical methods for detector calibration and operation.

    PubMed

    Borghi, Giacomo; Tabacchini, Valerio; Schaart, Dennis R

    2016-07-01

    Gamma-ray detectors based on thick monolithic scintillator crystals can achieve spatial resolutions  <2 mm full-width-at-half-maximum (FWHM) and coincidence resolving times (CRTs) better than 200 ps FWHM. Moreover, they provide high sensitivity and depth-of-interaction (DOI) information. While these are excellent characteristics for clinical time-of-flight (TOF) positron emission tomography (PET), the application of monolithic scintillators has so far been hampered by the lengthy and complex procedures needed for position- and time-of-interaction estimation. Here, the algorithms previously developed in our group are revised to make the calibration and operation of a large number of monolithic scintillator detectors in a TOF-PET system practical. In particular, the k-nearest neighbor (k-NN) classification method for x,y-position estimation is accelerated with an algorithm that quickly preselects only the most useful reference events, reducing the computation time for position estimation by a factor of ~200 compared to the previously published k-NN 1D method. Also, the procedures for estimating the DOI and time of interaction are revised to enable full detector calibration by means of fan-beam or flood irradiations only. Moreover, a new technique is presented to allow the use of events in which some of the photosensor pixel values and/or timestamps are missing (e.g. due to dead time), so as to further increase system sensitivity. The accelerated methods were tested on a monolithic scintillator detector specifically developed for clinical PET applications, consisting of a 32 mm  ×  32 mm  ×  22 mm LYSO : Ce crystal coupled to a digital photon counter (DPC) array. This resulted in a spatial resolution of 1.7 mm FWHM, an average DOI resolution of 3.7 mm FWHM, and a CRT of 214 ps. Moreover, the possibility of using events missing the information of up to 16 out of 64 photosensor pixels is shown. This results in only a small

  17. Towards monolithic scintillator based TOF-PET systems: practical methods for detector calibration and operation

    NASA Astrophysics Data System (ADS)

    Borghi, Giacomo; Tabacchini, Valerio; Schaart, Dennis R.

    2016-07-01

    Gamma-ray detectors based on thick monolithic scintillator crystals can achieve spatial resolutions  <2 mm full-width-at-half-maximum (FWHM) and coincidence resolving times (CRTs) better than 200 ps FWHM. Moreover, they provide high sensitivity and depth-of-interaction (DOI) information. While these are excellent characteristics for clinical time-of-flight (TOF) positron emission tomography (PET), the application of monolithic scintillators has so far been hampered by the lengthy and complex procedures needed for position- and time-of-interaction estimation. Here, the algorithms previously developed in our group are revised to make the calibration and operation of a large number of monolithic scintillator detectors in a TOF-PET system practical. In particular, the k-nearest neighbor (k-NN) classification method for x,y-position estimation is accelerated with an algorithm that quickly preselects only the most useful reference events, reducing the computation time for position estimation by a factor of ~200 compared to the previously published k-NN 1D method. Also, the procedures for estimating the DOI and time of interaction are revised to enable full detector calibration by means of fan-beam or flood irradiations only. Moreover, a new technique is presented to allow the use of events in which some of the photosensor pixel values and/or timestamps are missing (e.g. due to dead time), so as to further increase system sensitivity. The accelerated methods were tested on a monolithic scintillator detector specifically developed for clinical PET applications, consisting of a 32 mm  ×  32 mm  ×  22 mm LYSO : Ce crystal coupled to a digital photon counter (DPC) array. This resulted in a spatial resolution of 1.7 mm FWHM, an average DOI resolution of 3.7 mm FWHM, and a CRT of 214 ps. Moreover, the possibility of using events missing the information of up to 16 out of 64 photosensor pixels is shown. This results in only a small

  18. Balancing Particle Diversity in Markov Chain Monte Carlo Methods for Dual Calibration-Data Assimilation Problems in Hydrologic Modeling

    NASA Astrophysics Data System (ADS)

    Hernandez, F.; Liang, X.

    2014-12-01

    Given the inherent uncertainty in almost all of the variables involved, recent research is re-addressing the problem of calibrating hydrologic models from a stochastic perspective: the focus is shifting from finding a single parameter configuration that minimizes the model error, to approximating the maximum likelihood multivariate probability distribution of the parameters. To this end, Markov chain Monte Carlo (MCMC) formulations are widely used, where the distribution is defined as a smoothed ensemble of particles or members, each of which represents a feasible parameterization. However, the updating of these ensembles needs to strike a careful balance so that the particles adequately resemble the real distribution without either clustering or drifting excessively. In this study, we explore the implementation of two techniques that attempt to improve the quality of the resulting ensembles, both for the approximation of the model parameters and of the unknown states, in a dual calibration-data assimilation framework. The first feature of our proposed algorithm, in an effort to keep members from clustering on areas of high likelihood in light of the observations, is the introduction of diversity-inducing operators after each resampling. This approach has been successfully used before, and here we aim at testing additional operators which are also borrowed from the Evolutionary Computation literature. The second feature is a novel arrangement of the particles into two alternative data structures. The first one is a non-sorted Pareto population which favors 1) particles with high likelihood, and 2) particles that introduce a certain level of heterogeneity. The second structure is a partitioned array, in which each partition requires its members to have increasing levels of divergence from the particles in the areas of larger likelihood. Our newly proposed algorithm will be evaluated and compared to traditional MCMC methods in terms of convergence speed, and the

  19. Application of an Artificial Intelligence Method for Velocity Calibration and Events Location in Microseismic Monitoring

    NASA Astrophysics Data System (ADS)

    Yang, Y.; Chen, X.

    2013-12-01

    Good quality hydraulic fracture maps are heavily dependent upon the best possible velocity structure. Particle Swarm Optimization inversion scheme, an artificial intelligence technique for velocity calibration and events location could serve as a viable option, able to produce high quality data. Using perforation data to recalibrate the 1D isotropic velocity model derived from dipole sonic logs (or even without them), we are able to get the initial velocity model used for consequential events location. Velocity parameters can be inverted, as well as the thickness of the layer, through an iterative procedure. Performing inversion without integrating available data is unlikely to produce reliable results; especially if there are only one perforation shot and a single poor-layer-covered array along with low signal/noise ratio signal. The inversion method was validated via simulations and compared to the Fast Simulated Annealing approach and the Conjugate Gradient method. Further velocity model refinement can be accomplished while calculating events location during the iterative procedure minimizing the residuals from both sides. This artificial intelligence technique also displays promising application to the joint inversion of large-scale seismic activities data.

  20. Method for calibrating a Fourier transform ion cyclotron resonance mass spectrometer

    DOEpatents

    Smith, Richard D.; Masselon, Christophe D.; Tolmachev, Aleksey

    2003-08-19

    A method for improving the calibration of a Fourier transform ion cyclotron resonance mass spectrometer wherein the frequency spectrum of a sample has been measured and the frequency (f) and intensity (I) of at least three species having known mass to charge (m/z) ratios and one specie having an unknown (m/z) ratio have been identified. The method uses the known (m/z) ratios, frequencies, and intensities at least three species to calculate coefficients A, B, and C, wherein the mass to charge ratio of a least one of the three species (m/z).sub.i is equal to ##EQU1## wherein f.sub.i is the detected frequency of the specie, G(I.sub.i) is a predetermined function of the intensity of the species, and Q is a predetermined exponent. Using the calculated values for A, B, and C, the mass to charge ratio of the unknown specie (m/z).sub.ii is calculated as the sum of ##EQU2## wherein f.sub.ii is the measured frequency of the unknown specie, and (I.sub.ii) is the measured intensity of the unknown specie.

  1. Satellite-Sensor Calibration Verification Using the Cloud-Shadow Method

    NASA Technical Reports Server (NTRS)

    Reinersman, P.; Carder, K. L.; Chen, F. R.

    1995-01-01

    An atmospheric-correction method which uses cloud-shaded pixels together with pixels in a neighboring region of similar optical properties is described. This cloud-shadow method uses the difference between the total radiance values observed at the sensor for these two regions, thus removing the nearly identical atmospheric radiance contributions to the two signals (e.g. path radiance and Fresnel-reflected skylight). What remains is largely due to solar photons backscattered from beneath the sea to dominate the residual signal. Normalization by the direct solar irradiance reaching the sea surface and correction for some second-order effects provides the remote-sensing reflectance of the ocean at the location of the neighbor region, providing a known 'ground target' spectrum for use in testing the calibration of the sensor. A similar approach may be useful for land targets if horizontal homogeneity of scene reflectance exists about the shadow. Monte Carlo calculations have been used to correct for adjacency effects and to estimate the differences in the skylight reaching the shadowed and neighbor pixels.

  2. a Novel Approach to Camera Calibration Method for Smart Phones Under Road Environment

    NASA Astrophysics Data System (ADS)

    Lee, Bijun; Zhou, Jian; Ye, Maosheng; Guo, Yuan

    2016-06-01

    Monocular vision-based lane departure warning system has been increasingly used in advanced driver assistance systems (ADAS). By the use of the lane mark detection and identification, we proposed an automatic and efficient camera calibration method for smart phones. At first, we can detect the lane marker feature in a perspective space and calculate edges of lane markers in image sequences. Second, because of the width of lane marker and road lane is fixed under the standard structural road environment, we can automatically build a transformation matrix between perspective space and 3D space and get a local map in vehicle coordinate system. In order to verify the validity of this method, we installed a smart phone in the `Tuzhi' self-driving car of Wuhan University and recorded more than 100km image data on the road in Wuhan. According to the result, we can calculate the positions of lane markers which are accurate enough for the self-driving car to run smoothly on the road.

  3. A Novel Non-Intrusive Method to Resolve the Thermal-Dome-Effect of Pyranometers: Radiometric Calibration and Implications

    NASA Technical Reports Server (NTRS)

    Ji, Qiang; Tsay, Si-Chee; Lau, K. M.; Hansell, R. A.; Butler, J. J.; Cooper, J. W.

    2011-01-01

    Traditionally the calibration equation for pyranometers assumes that the measured solar irradiance is solely proportional to the thermopile's output voltage; therefore only a single calibration factor is derived. This causes additional measurement uncertainties because it does not capture sufficient information to correctly account for a pyranometer's thermal effect. In our updated calibration equation, temperatures from the pyranometer's dome and case are incorporated to describe the instrument's thermal behavior, and a new set of calibration constants are determined, thereby reducing measurement uncertainties. In this paper, we demonstrate why a pyranometer's uncertainty using the traditional calibration equation is always larger than a-few-percent, but with the new approach can become much less than 1% after the thermal issue is resolved. The highlighted calibration results are based on NIST-traceable light sources under controlled laboratory conditions. The significance of the new approach lends itself to not only avoiding the uncertainty caused by a pyranometer's thermal effect but also the opportunity to better isolate and characterize other instrumental artifacts, such as angular response and non-linearity of the thermopile, to further reduce additional uncertainties. We also discuss some of the implications, including an example of how the thermal issue can potentially impact climate studies by evaluating aerosol's direct-radiative effect using field measurements with and without considering the pyranometer's thermal effect. The results of radiative transfer model simulation show that a pyranometer's thermal effect on solar irradiance measurements at the surface can be translated into a significant alteration of the calculated distribution of solar energy inside the column atmosphere.

  4. Assessing calibration of prognostic risk scores.

    PubMed

    Crowson, Cynthia S; Atkinson, Elizabeth J; Therneau, Terry M

    2016-08-01

    Current methods used to assess calibration are limited, particularly in the assessment of prognostic models. Methods for testing and visualizing calibration (e.g. the Hosmer-Lemeshow test and calibration slope) have been well thought out in the binary regression setting. However, extension of these methods to Cox models is less well known and could be improved. We describe a model-based framework for the assessment of calibration in the binary setting that provides natural extensions to the survival data setting. We show that Poisson regression models can be used to easily assess calibration in prognostic models. In addition, we show that a calibration test suggested for use in survival data has poor performance. Finally, we apply these methods to the problem of external validation of a risk score developed for the general population when assessed in a special patient population (i.e. patients with particular comorbidities, such as rheumatoid arthritis). PMID:23907781

  5. Methods of travel-time residual declustering for the knowledge base calibration and integration tool (KBCIT)

    SciTech Connect

    Myers, S C

    2001-02-05

    Calibration of seismic nuclear test monitoring stations relies on a diverse reference-event database. The reference-event covariance structure must be characterized and the covariance propagated to subsequent processes and calibration products. In seismic location one of the first steps in the calibration process is declustering, in which closely spaced epicenters are combined to reduce redundant data and random observational errors. We formulate a new declustering procedure that accounts for correlated and uncorrelated components of reference-event error, producing a declustered data set that tracks raw reference-event uncertainties. Declustering behavior is demonstrated using example data sets.

  6. The Role of Mathematical Methods in Efficiency Calibration and Uncertainty Estimation in Gamma Based Non-Destructive Assay - 12311

    SciTech Connect

    Venkataraman, R.; Nakazawa, D.

    2012-07-01

    Mathematical methods are being increasingly employed in the efficiency calibration of gamma based systems for non-destructive assay (NDA) of radioactive waste and for the estimation of the Total Measurement Uncertainty (TMU). Recently, ASTM (American Society for Testing and Materials) released a standard guide for use of modeling passive gamma measurements. This is a testimony to the common use and increasing acceptance of mathematical techniques in the calibration and characterization of NDA systems. Mathematical methods offer flexibility and cost savings in terms of rapidly incorporating calibrations for multiple container types, geometries, and matrix types in a new waste assay system or a system that may already be operational. Mathematical methods are also useful in modeling heterogeneous matrices and non-uniform activity distributions. In compliance with good practice, if a computational method is used in waste assay (or in any other radiological application), it must be validated or benchmarked using representative measurements. In this paper, applications involving mathematical methods in gamma based NDA systems are discussed with several examples. The application examples are from NDA systems that were recently calibrated and performance tested. Measurement based verification results are presented. Mathematical methods play an important role in the efficiency calibration of gamma based NDA systems. This is especially true when the measurement program involves a wide variety of complex item geometries and matrix combinations for which the development of physical standards may be impractical. Mathematical methods offer a cost effective means to perform TMU campaigns. Good practice demands that all mathematical estimates be benchmarked and validated using representative sets of measurements. (authors)

  7. PLEIADES-HR 1A&1B image quality commissioning: innovative radiometric calibration methods and results

    NASA Astrophysics Data System (ADS)

    Martin, Vincent; Blanchet, Gwendoline; Kubik, Philippe; Lacherade, Sophie; Latry, Christophe; Lebegue, Laurent; Lenoir, Florie; Porez-Nadal, Florence

    2013-09-01

    PLEIADES is an earth observing system conducted by the French National Space Agency, CNES. It consists of two satellites launched on December 2011 (PHR-1A) and December 2012 (PHR-1B), both designed to provide optical pushbroom imagery on five spectral bands to civilian and defense users, with ground sample distance up to 70 cm. During inflight image quality commissioning, radiometric activities included inter-detector normalization coefficients computation, refocusing operations, MTF assessment and estimation of signal to noise ratios. This paper presents inflight results for both satellites. It focuses on several innovative methods that were implemented, taking advantage of the satellite platform great agility. These methods are based on processing images obtained through dedicated exotic guidance. In particular, slow-motion steering enables an efficient estimation of the instrumental noise model, since during acquisition each detector has been viewing a stable ground target along different time samples. Conversely, rotated retina guidance is used to guarantee that all different elementary detectors have successively viewed the same set of landscape samples during acquisition. Non-uniformity of detector sensitivities can then be characterized, and on-board coefficients used prior to compression can be calibrated in order to prevent vertical striping effects on operational images. Defocus control and Point Spread Function estimation can be easily obtained through processing acquisitions of stars associated to various spectral characteristics, for different adjustments of the refocusing system. All these methods allow an accurate estimation of radiometric performance on the whole range of specified spectral radiances, while drastically reducing the number of required acquisitions on natural targets.

  8. Study on method of radiometric calibration for precision measurement of micro size damage site

    NASA Astrophysics Data System (ADS)

    Yuan, Hao-yu; Peng, Zhi-tao; Wang, Wen-fang; Chen, Feng-dong; Tang, Jun; Feng, Bo; Liu, Guo-dong; Liu, Bing-guo

    2014-09-01

    Large aperture optical have high risk of damage when woke on high flux laser. For avoid lethal damages breakdown the expensive large aperture optical, replace the optical that damaged before damage site increase to can't repaired, we need precision measurement of optical surface damage sites size. The size of the optics which be detected is 400μm ×400μm, and the size of CCD array pixel is 4K×4K which we selected, so pixel resolution only 100μm of the Optical Damage Online Inspection system, it hard to measurement damage sites which size less than 100μm. This paper describes a method of radiometric calibration to measure online optical damage site that greater than 50μm by Optical Damage Online Inspection system. Numerical statement gray on CCD of different size damage sites by select a fixed variable of illumination intensity, shutter and numerical aperture of image-forming system. Fitting a curve with suitable function of gray and actual size, precision measure optical damage sites that greater than 50μm by the curve. Test results indicate that, the deviation less than 20% which measure size and actual size .This method settle problems of micro size damage site hard to measure online under the condition of long working distance and low optical resolution. At present, this method have used on Optical Damage Online Inspection system of high flux laser installation, it important significance for observation damage site size grown and accurately appraise the optical damage.

  9. Passive sampling methods for contaminated sediments: Practical guidance for selection, calibration, and implementation

    PubMed Central

    Ghosh, Upal; Driscoll, Susan Kane; Burgess, Robert M; Jonker, Michiel To; Reible, Danny; Gobas, Frank; Choi, Yongju; Apitz, Sabine E; Maruya, Keith A; Gala, William R; Mortimer, Munro; Beegan, Chris

    2014-01-01

    This article provides practical guidance on the use of passive sampling methods (PSMs) that target the freely dissolved concentration (Cfree) for improved exposure assessment of hydrophobic organic chemicals in sediments. Primary considerations for selecting a PSM for a specific application include clear delineation of measurement goals for Cfree, whether laboratory-based “ex situ” and/or field-based “in situ” application is desired, and ultimately which PSM is best-suited to fulfill the measurement objectives. Guidelines for proper calibration and validation of PSMs, including use of provisional values for polymer–water partition coefficients, determination of equilibrium status, and confirmation of nondepletive measurement conditions are defined. A hypothetical example is described to illustrate how the measurement of Cfree afforded by PSMs reduces uncertainty in assessing narcotic toxicity for sediments contaminated with polycyclic aromatic hydrocarbons. The article concludes with a discussion of future research that will improve the quality and robustness of Cfree measurements using PSMs, providing a sound scientific basis to support risk assessment and contaminated sediment management decisions. Integr Environ Assess Manag 2014;10:210–223. © 2014 The Authors. Integrated Environmental Assessment and Management published by Wiley Periodicals, Inc. on behalf of SETAC. PMID:24288273

  10. Light calibration and quality assessment methods for Reflectance Transformation Imaging applied to artworks' analysis

    NASA Astrophysics Data System (ADS)

    Giachetti, A.; Daffara, C.; Reghelin, C.; Gobbetti, E.; Pintus, R.

    2015-06-01

    In this paper we analyze some problems related to the acquisition of multiple illumination images for Polynomial Texture Maps (PTM) or generic Reflectance Transform Imaging (RTI). We show that intensity and directionality nonuniformity can be a relevant issue when acquiring manual sets of images with the standard highlight-based setup both using a flash lamp and a LED light. To maintain a cheap and flexible acquisition setup that can be used on field and by non-experienced users we propose to use a dynamic calibration and correction of the lights based on multiple intensity and direction estimation around the imaged object during the acquisition. Preliminary tests on the results obtained have been performed by acquiring a specifically designed 3D printed pattern in order to see the accuracy of the acquisition obtained both for spatial discrimination of small structures and normal estimation, and on samples of different types of paper in order to evaluate material discrimination. We plan to design and build from our analysis and from the tools developed and under development a set of novel procedures and guidelines that can be used to turn the cheap and common RTI acquisition setup from a simple way to enrich object visualization into a powerful method for extracting quantitative characterization both of surface geometry and of reflective properties of different materials. These results could have relevant applications in the Cultural Heritage domain, in order to recognize different materials used in paintings or investigate the ageing status of artifacts' surface.

  11. Pick and Choose the Spectroscopic Method to Calibrate the Local Electric Field inside Proteins.

    PubMed

    Haldar, Tapas; Kashid, Somnath M; Deb, Pranab; Kesh, Sandeep; Bagchi, Sayan

    2016-07-01

    Electrostatic interactions in proteins play a crucial role in determining the structure-function relation in biomolecules. In recent years, fluorescent probes have been extensively employed to interrogate the polarity in biological cavities through dielectric constants or semiempirical polarity scales. A choice of multiple spectroscopic methods, not limited by fluorophores, along with a molecular level description of electrostatics involving solute-solvent interactions, would allow more flexibility to pick and choose the experimental technique to determine the local electrostatics within protein interiors. In this work we report that ultraviolet/visible-absorption, infrared-absorption, or (13)C NMR can be used to calibrate the local electric field in both hydrogen bonded and non-hydrogen bonded protein environments. The local electric field at the binding site of a serum protein has been determined using the absorption wavelength as well as the carbonyl stretching frequency of its natural steroid substrate, testosterone. Excellent agreement is observed in the results obtained from two independent spectroscopic techniques. PMID:27295386

  12. Calibrating a method for simulated long-term ageing of biochar

    NASA Astrophysics Data System (ADS)

    Sohi, Saran; Cross, Andrew

    2013-04-01

    We recently established a procedure that imposes oxidatiave ageing to biochar and charcoal samples over a short time-frame, that provided carbon mass loss in the range projected for wild-fire charcoal in soil over a period of approximately 100 years. The stability of biochar samples in soil (relative to charcoal) range from 45-98% could be determined repeatably with high precision. Initial tests to understand the kinetics of the accelerated ageing method showed progressive increase in surface O concentration when examined by X-ray photoelectron spectroscopy (XPS) that slowly reached equilibrium. These trends resembled patterns observed in climate-for-time studies elsewhere, on centennial time-frame. We have extended this work to a preliminary direct calibration by matching progressive oxidation achieved in the laboratory to the surface composition of charcoal fragments recovered from the environment after periods of hundred to thousands of years. We have also applied artificial ageing to the same sets of naturally pre-aged charcoal fragments, and to recreated fresh charcoal. In this presentation of the first approach to quantifiably relate a laboratory test for biochar carbon stability to field data covering multiple time scales, we report on both the process and the implications for the stability of carbon stored in biochar under different climates and diverse agro-ecosystems.

  13. Research on autofocusing method with automatic calibration for aerial camera based on imaging resolution

    NASA Astrophysics Data System (ADS)

    Zhao, Yu-liang; Zhao, Hong-qiang; Li, Shu-jun; Zhang, Yu-ye

    2014-09-01

    Air materiel depot is a warehouse which store consumed all the parts and equipment vault of the plane. In order to ensure the various aviation equipment integrity of the backup piece rate, the inside temperature of depot must be controlled within a certain range. Therefore, the depot must be equipped a self-contained temperature real-time monitoring system. This paper presents a distributed temperature sensing alarm system to apply to real-time measure spatial distribution of temperature field. In order to eliminate influence to the scattering strength from the light source instability and the fiber bending splice loss and to improve temperature measurement accuracy, the system design used dual-channel dual-wavelength comparison method which make Anti-Stokes as signal channel and Stokes as a reference channel to collect signals of two channel respectively and detect the ratio of the two channels' signals. The light of LD directional coupling to the sensing optical fiber in the temperature field to test, domain reflect light from the sensing optical fiber directional coupling to receive channel again, Rayleigh domain reflect light is filtered after optical filter, the Anti-Stokes and Stokes are both taken out, converted and magnified, the two signals is digitalized by A/D Converter, and written to the storage machine , which linear cumulative to the content of the storage unit, The distributed measurement of the temperature field to test is finished. The collected 2900 measuring points real-time on 2km of optical fiber. The spatial resolution of the system was 0.7m, measurement range was -20-370 °C, and measurement error was +/- 2 °C. All index of the system achieved the desired objective. To get an accurate temperature field spatial distribution and the information of temporal variation, the system enabled real-time temperature of aviation depot monitoring and early warning. As a new sensing technology, the distributed fiber optic sensor has the functions of self

  14. Calibration of X-Ray diffractometer by the experimental comparison method

    NASA Astrophysics Data System (ADS)

    Dudka, A. P.

    2015-07-01

    A software for calibrating an X-ray diffractometer with area detector has been developed. It is proposed to search for detector and goniometer calibration models whose parameters are reproduced in a series of measurements on a reference crystal. Reference (standard) crystals are prepared during the investigation; they should provide the agreement of structural models in repeated analyses. The technique developed has been used to calibrate Xcalibur Sapphire and Eos, Gemini Ruby (Agilent) and Apex x8 and Apex Duo (Bruker) diffractometers. The main conclusions are as follows: the calibration maps are stable for several years and can be used to improve structural results, verified CCD detectors exhibit significant inhomogeneity of the efficiency (response) function, and a Bruker goniometer introduces smaller distortions than an Agilent goniometer.

  15. Calibration of X-Ray diffractometer by the experimental comparison method

    SciTech Connect

    Dudka, A. P.

    2015-07-15

    A software for calibrating an X-ray diffractometer with area detector has been developed. It is proposed to search for detector and goniometer calibration models whose parameters are reproduced in a series of measurements on a reference crystal. Reference (standard) crystals are prepared during the investigation; they should provide the agreement of structural models in repeated analyses. The technique developed has been used to calibrate Xcalibur Sapphire and Eos, Gemini Ruby (Agilent) and Apex x8 and Apex Duo (Bruker) diffractometers. The main conclusions are as follows: the calibration maps are stable for several years and can be used to improve structural results, verified CCD detectors exhibit significant inhomogeneity of the efficiency (response) function, and a Bruker goniometer introduces smaller distortions than an Agilent goniometer.

  16. Calibration of sound calibrators: an overview

    NASA Astrophysics Data System (ADS)

    Milhomem, T. A. B.; Soares, Z. M. D.

    2016-07-01

    This paper presents an overview of calibration of sound calibrators. Initially, traditional calibration methods are presented. Following, the international standard IEC 60942 is discussed emphasizing parameters, target measurement uncertainty and criteria for conformance to the requirements of the standard. Last, Regional Metrology Organizations comparisons are summarized.

  17. The use of suitable pseudo-invariant targets for MIVIS data calibration by the empirical line method

    NASA Astrophysics Data System (ADS)

    Mei, Alessandro; Bassani, Cristiana; Fontinovo, Giuliano; Salvatori, Rosamaria; Allegrini, Alessia

    2016-04-01

    The Empirical Line Method (ELM) enables the calibration of multi- and hyper-airborne/satellite image converting DN or radiance to reflectance values performed by using at ground data. High quality outcome can be reached with the selection of appropriate Pseudo-Invariant Targets (PIT). In this paper, spectral variability of "usual" (asphalt and concrete) and "unusual" (calcareous gravel, basaltic paving, concrete bricks, tartan paving and artificial turf) PITs is retrieved for ELM application. Such PITs are used to calibrate the Multispectral Infrared and Visible Imaging Spectrometer (MIVIS) airborne sensor in 12 different Runs. Firstly, processing of field spectral data enables the evaluation of pseudo-invariance of targets by studying their spectral changes in space and in time. Finally, these surfaces are used as Ground Calibration (GCT) and Validation Targets (GVT) in ELM. High calibration accuracy values are observed in Visible (VIS) range (98.9%) while a general decrease of accuracy in Near-InfraRed (NIR) (96.6%) and Middle-InfraRed (SWIR) (88.1%) are reached. Outcomes show that "usual" surfaces as asphalt and concrete and "unusual" surfaces such as tartan can be successfully used for MIVIS image calibration.

  18. A Monte Carlo (MC) based individual calibration method for in vivo x-ray fluorescence analysis (XRF)

    NASA Astrophysics Data System (ADS)

    Hansson, Marie; Isaksson, Mats

    2007-04-01

    X-ray fluorescence analysis (XRF) is a non-invasive method that can be used for in vivo determination of thyroid iodine content. System calibrations with phantoms resembling the neck may give misleading results in the cases when the measurement situation largely differs from the calibration situation. In such cases, Monte Carlo (MC) simulations offer a possibility of improving the calibration by better accounting for individual features of the measured subjects. This study investigates the prospects of implementing MC simulations in a calibration procedure applicable to in vivo XRF measurements. Simulations were performed with Penelope 2005 to examine a procedure where a parameter, independent of the iodine concentration, was used to get an estimate of the expected detector signal if the thyroid had been measured outside the neck. An attempt to increase the simulation speed and reduce the variance by exclusion of electrons and by implementation of interaction forcing was conducted. Special attention was given to the geometry features: analysed volume, source-sample-detector distances, thyroid lobe size and position in the neck. Implementation of interaction forcing and exclusion of electrons had no obvious adverse effect on the quotients while the simulation time involved in an individual calibration was low enough to be clinically feasible.

  19. Calibration method of laser plane equation for vision measurement adopting objective function of uniform horizontal height of feature points

    NASA Astrophysics Data System (ADS)

    Xu, Guan; Hao, Zhaobing; Li, Xiaotao; Su, Jian; Liu, Huanping; Zhang, Xinyuan

    2016-02-01

    A calibration method with an objective function generated from a uniform horizontal height is presented in this work for the laser plane in active vision measurement. A height target is developed with a center mark as the initial point of the uniform height. The height target is located on the horizontal plane of the 3D calibration board so that the horizontal plane is considered as the terminal of the uniform horizontal height. Based on the pinhole model of the camera and the laser plane equation, we model the objective function to find the optimal coefficients of the laser plane equation. The goal of the objective function is the smallest difference of the uniform height and the reconstructed height according to the feature points of the target. The objective function is optimized by the local particle swarm optimization. The calibrated global equation of a laser plane is obtained from the optimal value 1.153 × 103 of the objective function in the experiments. Two projective laser lines of the calibration laser plane cover the original laser lines in the image. The reconstruction errors of the calibration plane are also analyzed in discussions.

  20. Primary calibration of solar cells based on DSR method at the National Institute of Metrology of China

    NASA Astrophysics Data System (ADS)

    He, Yingwei; Xiong, Limin; Zhang, Junchao; Meng, Haifeng; Cai, Chuan; Zhang, Bifeng; Xie, Linlin; Liu, Dingpu

    2015-08-01

    A primary standard measurement facility based on differential spectral responsivity (DSR) method for calibration of reference solar cells was realized at National Institute of Metrology (NIM), China. The primary calibration of the critical spectral parameters and short-circuit current of reference cells, not only with WPVS (World photovoltaic Scale) design but with non-regularly shaped, can be performed by this standard facility. The linearity measurement can be carried out by measuring DSR of the solar cells at different bias levels in the spectral range from 300nm to 1200nm. The characterization and performance of the facility were reported. An uncertainty of 0.9% (k=2) for short-circuit current of WPVS reference solar cells was able to be obtained. A more accurate and better calibration service for solar photovoltaic (PV) cells could be provided to local or international solar cell research community, testing labs and industry users and manufacturers.

  1. Technique for Radiometer and Antenna Array Calibration with a Radiated Noise Diode

    NASA Technical Reports Server (NTRS)

    Srinivasan, Karthik; Limaye, Ashutosh; Laymon, Charles; Meyer, Paul

    2009-01-01

    This paper presents a new technique to calibrate a microwave radiometer and antenna array system. This calibration technique uses a radiated noise source in addition to two calibration sources internal to the radiometer. The method accurately calibrates antenna arrays with embedded active devices (such as amplifiers) which are used extensively in active phased array antennas.

  2. Key wavelengths screening using competitive adaptive reweighted sampling method for multivariate calibration.

    PubMed

    Li, Hongdong; Liang, Yizeng; Xu, Qingsong; Cao, Dongsheng

    2009-08-19

    By employing the simple but effective principle 'survival of the fittest' on which Darwin's Evolution Theory is based, a novel strategy for selecting an optimal combination of key wavelengths of multi-component spectral data, named competitive adaptive reweighted sampling (CARS), is developed. Key wavelengths are defined as the wavelengths with large absolute coefficients in a multivariate linear regression model, such as partial least squares (PLS). In the present work, the absolute values of regression coefficients of PLS model are used as an index for evaluating the importance of each wavelength. Then, based on the importance level of each wavelength, CARS sequentially selects N subsets of wavelengths from N Monte Carlo (MC) sampling runs in an iterative and competitive manner. In each sampling run, a fixed ratio (e.g. 80%) of samples is first randomly selected to establish a calibration model. Next, based on the regression coefficients, a two-step procedure including exponentially decreasing function (EDF) based enforced wavelength selection and adaptive reweighted sampling (ARS) based competitive wavelength selection is adopted to select the key wavelengths. Finally, cross validation (CV) is applied to choose the subset with the lowest root mean square error of CV (RMSECV). The performance of the proposed procedure is evaluated using one simulated dataset together with one near infrared dataset of two properties. The results reveal an outstanding characteristic of CARS that it can usually locate an optimal combination of some key wavelengths which are interpretable to the chemical property of interest. Additionally, our study shows that better prediction is obtained by CARS when compared to full spectrum PLS modeling, Monte Carlo uninformative variable elimination (MC-UVE) and moving window partial least squares regression (MWPLSR). PMID:19616692

  3. Traceability of Acoustic Emission measurements for a proposed calibration method - Classification of characteristics and identification using signal analysis

    NASA Astrophysics Data System (ADS)

    Griffin, James

    2015-01-01

    When using Acoustic Emission (AE) technologies, tensile, compressive and shear stress/strain tests can provide a detector for material deformation and dislocations. In this paper improvements are made to standardise calibration techniques for AE against known metrics such as force. AE signatures were evaluated from various calibration energy sources based on the energy from the first harmonic (dominant energy band) [1,2]. The effects of AE against its calibration identity are investigated: where signals are correlated to the average energy and distance of the detected phenomena. In addition, extra tests are investigated in terms of the tensile tests and single grit tests characterising different materials. Necessary translations to the time-frequency domain were necessary when segregating salient features between different material properties. Continuing this work the obtained AE is summarised and evaluated by a Neural Network (NN) regression classification technique which identifies how far the malformation has progressed (in terms of energy/force) during material transformation. Both genetic-fuzzy clustering and tree rule based classifier techniques were used as the second and third classification techniques respectively to verify the NN output giving a weighted three classifier system. The work discussed in this paper looks at both distance and force relationships for various prolonged Acoustic Emission stresses. Later such analysis was realised with different classifier models and finally implemented into the Simulink simulations. Further investigations were made into classifier models for different material interactions in terms of force and distance which add further dimension to this work with different materials based simulation realisations. Within the statistical analysis section there are two varying prolonged stress tests which together offer the mechanical calibration system (automated solenoid and pencil break calibration system). Taking such a

  4. The extended wedge method: Atomic force microscope friction calibration for improved tolerance to instrument misalignments, tip offset, and blunt probes

    NASA Astrophysics Data System (ADS)

    Khare, H. S.; Burris, D. L.

    2013-05-01

    One of the major challenges in understanding and controlling friction is the difficulty in bridging the length and time scales of macroscale contacts and those of the single asperity interactions they comprise. While the atomic force microscope (AFM) offers a unique ability to probe tribological surfaces in a wear-free single-asperity contact, instrument calibration challenges have limited the usefulness of this technique for quantitative nanotribological studies. A number of lateral force calibration techniques have been proposed and used, but none has gained universal acceptance due to practical considerations, configuration limitations, or sensitivities to unknowable error sources. This paper describes a simple extension of the classic wedge method of AFM lateral force calibration which: (1) allows simultaneous calibration and measurement on any substrate, thus eliminating prior tip damage and confounding effects of instrument setup adjustments; (2) is insensitive to adhesion, PSD cross-talk, transducer/piezo-tube axis misalignment, and shear-center offset; (3) is applicable to integrated tips and colloidal probes; and (4) is generally applicable to any reciprocating friction coefficient measurement. The method was applied to AFM measurements of polished carbon (99.999% graphite) and single crystal MoS2 to demonstrate the technique. Carbon and single crystal MoS2 had friction coefficients of μ = 0.20 ± 0.04 and μ = 0.006 ± 0.001, respectively, against an integrated Si probe. Against a glass colloidal sphere, MoS2 had a friction coefficient of μ = 0.005 ± 0.001. Generally, the measurement uncertainties ranged from 10%-20% and were driven by the effect of actual frictional variation on the calibration rather than calibration error itself (i.e., due to misalignment, tip-offset, or probe radius).

  5. A method for accurate zero calibration of asymmetric jaws in single-isocenter half-beam techniques

    SciTech Connect

    Hernandez, V.; Abella, R.; Lopez, M.; Perez, M.; Artigues, M.; Sempau, J.; Arenas, M.

    2013-02-15

    Purpose: To present a practical method for calibrating the zero position of asymmetric jaws that provides higher accuracy at the central axis and improves dose homogeneity in the abutting region of half-beams. Methods: Junction doses were measured for each asymmetric jaw using the double-exposure technique and electronic portal imaging devices. The junction dose was determined as a function of jaw position. The shift in the zero jaw position (or in its corresponding potentiometer readout) required to correct for the measured junction dose could thus be obtained. The jaw calibration was then modified to introduce the calculated shift and therefore achieve an accurate zero position in order to provide a relative junction dose that was as close to zero as possible. Results: All the asymmetric jaws from four medical linear accelerators were calibrated with the new calibration procedure. Measured relative junction doses at gantry 0 Degree-Sign were reduced from a maximum of {+-}40% to a maximum of {+-}8% for all the jaws in the four considered accelerators. These results were valid for 6 MV and 18 MV photon beams and for any combination of asymmetric jaws set to zero. The calibration was stable over a long period of time; therefore, the need for recalibrating is seldom necessary. Conclusions: Accurate calibration of the zero position of the jaws is feasible in current medical linear accelerators. The proposed procedure is fast and it improves dose homogeneity at the junction of half-beams, thus, allowing a more accurate and safer use of these techniques.

  6. Net analyte signal standard addition method for simultaneous determination of sulphadiazine and trimethoprim in bovine milk and veterinary medicines.

    PubMed

    Hajian, Reza; Mousavi, Esmat; Shams, Nafiseh

    2013-06-01

    Net analyte signal standard addition method has been used for the simultaneous determination of sulphadiazine and trimethoprim by spectrophotometry in some bovine milk and veterinary medicines. The method combines the advantages of standard addition method with the net analyte signal concept which enables the extraction of information concerning a certain analyte from spectra of multi-component mixtures. This method has some advantages such as the use of a full spectrum realisation, therefore it does not require calibration and prediction step and only a few measurements require for the determination. Cloud point extraction based on the phenomenon of solubilisation used for extraction of sulphadiazine and trimethoprim in bovine milk. It is based on the induction of micellar organised media by using Triton X-100 as an extraction solvent. At the optimum conditions, the norm of NAS vectors increased linearly with concentrations in the range of 1.0-150.0 μmolL(-1) for both sulphadiazine and trimethoprim. The limits of detection (LOD) for sulphadiazine and trimethoprim were 0.86 and 0.92 μmolL(-1), respectively. PMID:23411170

  7. A comparison of methods for calibration and use of multi-component strain gauge wind tunnel balances

    NASA Astrophysics Data System (ADS)

    Galway, R. D.

    1980-03-01

    A method is presented for calibration of strain-gauge balances which does not require that the components can be loaded independently. Applicable to both 'internal' and 'external' types of balance, the procedure uses a single varying calibration load to determine all linear and non-linear calibration coefficients. Constant 'secondary' loads on one or more components are unnecessary, although they may be used if desired. The usual iterative solution of the second order balance equations is outlined, and an approximate non-iterative scheme is included for completeness, though not recommended. Two methods of accounting for dependency of the calibration coefficients on the signs of the component loads are presented. A concept of 'buoyancy' is introduced to simplify the application of force balance tares, and a procedure for determining the component outputs for absolute zero load (the 'buoyant' offsets) is given. Balance data at a series of model attitudes are used to define these offsets, and also the coefficients in the equations defining the component load distribution of the tare weight at any attitude. The topics covered are ideally suited to formulation and solution by matrix methods, which have been used throughout.

  8. Joining direct and indirect inverse calibration methods to characterize karst, coastal aquifers

    NASA Astrophysics Data System (ADS)

    De Filippis, Giovanna; Foglia, Laura; Giudici, Mauro; Mehl, Steffen; Margiotta, Stefano; Negri, Sergio

    2016-04-01

    Parameter estimation is extremely relevant for accurate simulation of groundwater flow. Parameter values for models of large-scale catchments are usually derived from a limited set of field observations, which can rarely be obtained in a straightforward way from field tests or laboratory measurements on samples, due to a number of factors, including measurement errors and inadequate sampling density. Indeed, a wide gap exists between the local scale, at which most of the observations are taken, and the regional or basin scale, at which the planning and management decisions are usually made. For this reason, the use of geologic information and field data is generally made by zoning the parameter fields. However, pure zoning does not perform well in the case of fairly complex aquifers and this is particularly true for karst aquifers. In fact, the support of the hydraulic conductivity measured in the field is normally much smaller than the cell size of the numerical model, so it should be upscaled to a scale consistent with that of the numerical model discretization. Automatic inverse calibration is a valuable procedure to identify model parameter values by conditioning on observed, available data, limiting the subjective evaluations introduced with the trial-and-error technique. Many approaches have been proposed to solve the inverse problem. Generally speaking, inverse methods fall into two groups: direct and indirect methods. Direct methods allow determination of hydraulic conductivities from the groundwater flow equations which relate the conductivity and head fields. Indirect methods, instead, can handle any type of parameters, independently from the mathematical equations that govern the process, and condition parameter values and model construction on measurements of model output quantities, compared with the available observation data, through the minimization of an objective function. Both approaches have pros and cons, depending also on model complexity. For

  9. A likelihood method to cross-calibrate air-shower detectors

    NASA Astrophysics Data System (ADS)

    Dembinski, Hans Peter; Kégl, Balázs; Mariş, Ioana C.; Roth, Markus; Veberič, Darko

    2016-01-01

    We present a detailed statistical treatment of the energy calibration of hybrid air-shower detectors, which combine a surface detector array and a fluorescence detector, to obtain an unbiased estimate of the calibration curve. The special features of calibration data from air showers prevent unbiased results, if a standard least-squares fit is applied to the problem. We develop a general maximum-likelihood approach, based on the detailed statistical model, to solve the problem. Our approach was developed for the Pierre Auger Observatory, but the applied principles are general and can be transferred to other air-shower experiments, even to the cross-calibration of other observables. Since our general likelihood function is expensive to compute, we derive two approximations with significantly smaller computational cost. In the recent years both have been used to calibrate data of the Pierre Auger Observatory. We demonstrate that these approximations introduce negligible bias when they are applied to simulated toy experiments, which mimic realistic experimental conditions.

  10. Reproducible methods for calibrating the backscattered electron signal for quantitative assessment of mineral content in bone

    SciTech Connect

    Boyce, T.M.; Bloebaum, R.D.; Bachus, K.N.; Skedros, J.G. )

    1990-09-01

    Backscattered electron (BSE) imaging shows promise for orthopaedic and bone research. BSE images of bone may be captured on-line directly from the scanning electron microscope (SEM), and then analyzed to produce a backscattered electron profile (BSEP), a modified image graylevel histogram which is representative of the mineral content in bone. The goals of this work were (1) develop a reproducible graylevel calibration technique for bone specimens, and (2) determine a conservative time interval during which SEM operating conditions would remain stable. Calibration standards containing pure aluminum and pure magnesium wires were placed in the SEM with human cancellous bone. Baseline imaging conditions were first established by adjusting the SEM until the bone image displayed good resolution and graylevel separation between regions of different mineral content. Microscope brightness and contrast controls were randomly changed to initiate the new operating conditions of another imaging session, and graylevel values from the calibration metals were used to readjust the microscope back to baseline operating conditions. Weighted mean graylevel values of the BSEPs from calibration trials were compared to those of the baseline. Data showed that bone images could be reproduced within 1.2 percent. It was also concluded that our equipment required calibration checks at 20 minute intervals.

  11. A Method to Calibrate the High-resolution Catania Astrophysical Observatory Spectropolarimeter

    NASA Astrophysics Data System (ADS)

    Leone, F.; Avila, G.; Bellassai, G.; Bruno, P.; Catalano, S.; Di Benedetto, R.; Di Stefano, A.; Gangi, M.; Giarrusso, M.; Greco, V.; Martinetti, E.; Miraglia, M.; Munari, M.; Pontoni, C.; Scalia, C.; Scuderi, S.; Spanó, P.

    2016-05-01

    The Catania Astrophysical Observatory Spectropolarimeter (CAOS) is a white-pupil cross-dispersed échelle spectrograph with a spectral resolution of up to R = 55,000 in the 375–1100 nm range in a single exposure, with complete coverage up to 856 nm. CAOS is linked to the 36-inch telescope, at Mount Etna Observatory, with a couple of 100 μm optical fibers and it achieves a signal-to-noise ratio better than 60 for a V = 10 mag star in one hour. CAOS is thermally stabilized in temperature within a 0.01 K rms, so that radial velocities are measured with a precision better than 100 m s‑1 from a single spectral line. Linear and circular spectropolarimetric observations are possible by means of a Savart plate working in series with a half-wave and a quarter-wave retarder plate in the 376–850 nm range. As is usual for high-resolution spectropolarimeters, CAOS is suitable to measure all Stokes parameters across spectral lines and it cannot measure the absolute degree of polarization. Observations of unpolarized standard stars show that instrumental polarization is generally zero at 550 nm and can increase up to 3% at the other wavelengths. Since polarized and unpolarized standard stars are useless, we suggest a method to calibrate a high-resolution spectropolarimeter on the basis of the polarimetric properties of spectral lines formed in the presence of a magnetic field. As applied to CAOS, observations of magnetic chemically peculiar stars of the main sequence show that the cross-talk from linear to circular polarization is smaller than 0.4% and that conversion from circular to linear is less than 2.7%. Strength and wavelength dependences of cross-talk can be entirely ascribed, via numerical simulations, to the incorrect retardance of achromatic wave plates.

  12. Creating long term gridded fields of reference evapotranspiration in Alpine terrain based on a re-calibrated Hargreaves method

    NASA Astrophysics Data System (ADS)

    Haslinger, K.; Bartsch, A.

    2015-05-01

    A new approach for the construction of high resolution gridded fields of reference evapotranspiration for the Austrian domain on a daily time step is presented. Forcing fields of gridded data of minimum and maximum temperatures are used to estimate reference evapotranspiration based on the formulation of Hargreaves. The calibration constant in the Hargreaves equation is recalibrated to the Penman-Monteith equation, which is recommended by the FAO, in a monthly and station-wise assessment. This ensures on one hand eliminated biases of the Hargreaves approach compared to the formulation of Penman-Monteith and on the other hand also reduced root mean square errors and relative errors on a daily time scale. The resulting new calibration parameters are interpolated in time to a daily temporal resolution for a standard year of 365 days. The overall novelty of the approach is the conduction of surface elevation as a predictor to estimate the re-calibrated Hargreaves parameter in space. A third order spline is fitted to the re-calibrated parameters against elevation at every station and yields the statistical model for assessing these new parameters in space by using the underlying digital elevation model of the temperature fields. Having newly calibrated parameters for every day of year and every grid point, the Hargreaves method is applied to the temperature fields, yielding reference evapotranspiration for the entire grid and time period from 1961-2013. With this approach it is possible to generate high resolution reference evapotranspiration fields starting when only temperature observations are available but re-calibrated to meet the requirements of the recommendations defined by the FAO.

  13. A non-contact, thermal noise based method for the calibration of lateral deflection sensitivity in atomic force microscopy

    SciTech Connect

    Mullin, Nic Hobbs, Jamie K.

    2014-11-15

    Calibration of lateral forces and displacements has been a long standing problem in lateral force microscopies. Recently, it was shown by Wagner et al. that the thermal noise spectrum of the first torsional mode may be used to calibrate the deflection sensitivity of the detector. This method is quick, non-destructive and may be performed in situ in air or liquid. Here we make a full quantitative comparison of the lateral inverse optical lever sensitivity obtained by the lateral thermal noise method and the shape independent method developed by Anderson et al. We find that the thermal method provides accurate results for a wide variety of rectangular cantilevers, provided that the geometry of the cantilever is suitable for torsional stiffness calibration by the torsional Sader method, in-plane bending of the cantilever may be eliminated or accounted for and that any scaling of the lateral deflection signal between the measurement of the lateral thermal noise and the measurement of the lateral deflection is eliminated or corrected for. We also demonstrate that the thermal method may be used to characterize the linearity of the detector signal as a function of position, and find a deviation of less than 8% for the instrument used.

  14. Linear model correction: A method for transferring a near-infrared multivariate calibration model without standard samples.

    PubMed

    Liu, Yan; Cai, Wensheng; Shao, Xueguang

    2016-12-01

    Calibration transfer is essential for practical applications of near infrared (NIR) spectroscopy because the measurements of the spectra may be performed on different instruments and the difference between the instruments must be corrected. For most of calibration transfer methods, standard samples are necessary to construct the transfer model using the spectra of the samples measured on two instruments, named as master and slave instrument, respectively. In this work, a method named as linear model correction (LMC) is proposed for calibration transfer without standard samples. The method is based on the fact that, for the samples with similar physical and chemical properties, the spectra measured on different instruments are linearly correlated. The fact makes the coefficients of the linear models constructed by the spectra measured on different instruments are similar in profile. Therefore, by using the constrained optimization method, the coefficients of the master model can be transferred into that of the slave model with a few spectra measured on slave instrument. Two NIR datasets of corn and plant leaf samples measured with different instruments are used to test the performance of the method. The results show that, for both the datasets, the spectra can be correctly predicted using the transferred partial least squares (PLS) models. Because standard samples are not necessary in the method, it may be more useful in practical uses. PMID:27380302

  15. Calibration method for spectral responsivity of infrared detector based on blackbody at multiple temperature

    NASA Astrophysics Data System (ADS)

    Zhang, Y. F.; Shao, Z. F.; Wu, Y. Q.

    2015-08-01

    The spectral responsivity is one of the most important technical indicators of infrared detector which has an important significance for radiation thermometry and emissivity measurement. Using a blackbody radiation at multiple temperatures, the calibration for spectral responsivity of the infrared detector is proposed. With the Planck's law, the spectral radiance of blackbody at the different temperature is calculated. The detector captures the radiation and generates output values each of those is the function of spectral responsivity, spectral radiance and environmental radiation. Calibration equation is established by means of the calculated radiance and output values. By solving the equations based on principle of least squares, the calibration of spectral responsivity is implemented. From the comparison experiment of measuring the radiance of blackbody at 850K, radiance value measured by the MCT detector has a good consistency with the theoretical data.

  16. Rock magnetic properties of dusty olivine: comparison and calibration of non-heating paleointensity methods

    NASA Astrophysics Data System (ADS)

    Lappe, S. L.; Harrison, R. J.; Feinberg, J. M.

    2012-12-01

    The mechanism of chondrule formation is an important outstanding question in cosmochemistry. Magnetic signals recorded by Fe-Ni nanoparticles in chondrules could carry clues to their origin. Recently, research in this area has focused on 'dusty olivine' in ordinary chondrites as potential carriers of pre-accretionary remanence. Dusty olivine is characterised by the presence of sub-micron Fe-Ni inclusions within the olivine host. These metal particles form via subsolidus reduction of the olivine during chondrule formation and are thought to be protected from subsequent chemical and thermal alteration by the host olivine. Three sets of synthetic dusty olivines have been produced, using natural olivine (average Ni-content of 0.3 wt%), synthetic Ni-containing olivine (0.1wt% Ni) and synthetic Ni-free olivine as starting materials. The starting materials were ground to powders, packed into a 8-27 mm3 graphite crucible, heated up to 1350°C under a pure CO gas flow and kept at this temperature for 10 minutes. After this the samples were held in fixed orientation and quenched into water in a range of known magnetic fields from 0.2 mT to 1.5 mT. We present a comparison of all non-heating methods commonly used for paleointensity determination of extraterrestrial material. All samples showed uni-directional, single-component demagnetization behaviour. Saturation REM ratio (NRM/SIRM) and REMc ratio show non-linear behaviour as function of applied field and a saturation value < 1. Using the REM' method the samples showed approximately constant REM' between 100 and 150 mT AF-field. Plotting the average values for this field range again shows non-linear behaviour and a saturation value < 1. Another approach we examined to obtain calibration curves for paleointensity determination is based on ARM measurents. We also present an analysis of a new FORC-based method of paleointensity determination applied to metallic Fe-bearing samples [1, 2]. The method uses a first-order reversal

  17. A New Method for Electronic Recoil Calibration in Liquid Noble Dark Matter Detectors

    NASA Astrophysics Data System (ADS)

    Macmullin, Sean

    2014-03-01

    Calibration of next-generation liquid noble dark matter detectors present new challenges because radiation from external sources will not probe the entire target, owing to its large volume and high stopping power. For electronic recoil calibration in particular, a proposed solution is to dissolve a source of low-energy β-electrons directly into the liquid. A particularly promising candidate is 212Pb, a daughter of 220Rn. We have acquired a custom-made source of electrodeposited 228Th that efficiently emanates the desired 220Rn. Details of recent measurements of mixing 220Rn and its daughters in a liquid xenon detector and future prospects will be presented.

  18. New method of voxel phantom creation: application for whole-body counting calibration and perspectives in individual internal dose assessment.

    PubMed

    de Carlan, L; Roch, P; Blanchardon, E; Franck, D

    2005-01-01

    The purpose of this work is to present an innovative approach for the creation and application of voxel phantoms associated with the Monte Carlo calculation (MCNP) for the calibration of whole-body counting systems dedicated to the measurement of fission and activation products. The new method is based on a graphical user interface called 'OEDIPE' that allows to simulate a whole measurement process using all measurement parameters, the final goal being to approach a numerical calibration of the facilities. The creation of voxel phantoms and validation of the method are presented in this paper using the IGOR phantom. Finally, the efficiency of the method is discussed, in particular, with the perspective of validating IGOR as a suitable human-equivalent phantom and for the assessment of uncertainties in dose estimation due to the inhomogeneous distribution of activity in the body, correlated to the bio-kinetic behaviour of the radionuclides. PMID:16604619

  19. On the use of multi-algorithm, genetically adaptive multi-objective method for multi-site calibration of the SWAT model

    SciTech Connect

    Zhang, Xuesong; Srinivasan, Raghavan; Van Liew, M.

    2010-04-15

    With the availability of spatially distributed data, distributed hydrologic models are increasingly used for simulation of spatially varied hydrologic processes to understand and manage natural and human activities that affect watershed systems. Multi-objective optimization methods have been applied to calibrate distributed hydrologic models using observed data from multiple sites. As the time consumed by running these complex models is increasing substantially, selecting efficient and effective multi-objective optimization algorithms is becoming a nontrivial issue. In this study, we evaluated a multi-algorithm, genetically adaptive multi-objective method (AMALGAM) for multi-site calibration of a distributed hydrologic model—Soil and Water Assessment Tool (SWAT), and compared its performance with two widely used evolutionary multi-objective optimization (EMO) algorithms (i.e. Strength Pareto Evolutionary Algorithm 2 (SPEA2) and Non-dominated Sorted Genetic Algorithm II (NSGA-II)). In order to provide insights into each method’s overall performance, these three methods were tested in four watersheds with various characteristics. The test results indicate that the AMALGAM can consistently provide competitive or superior results compared with the other two methods. The multi-method search framework of AMALGAM, which can flexibly and adaptively utilize multiple optimization algorithms, makes it a promising tool for multi-site calibration of the distributed SWAT. For practical use of AMALGAM, it is suggested to implement this method in multiple trials with relatively small number of model runs rather than run it once with long iterations. In addition, incorporating different multiobjective optimization algorithms and multi-mode search operators into AMALGAM deserves further research.

  20. Calibration of the pressure sensitivity of microphones by a free-field method at frequencies up to 80 khz.

    PubMed

    Zuckerwar, Allan J; Herring, G C; Elbing, Brian R

    2006-01-01

    A free-field (FF) substitution method for calibrating the pressure sensitivity of microphones at frequencies up to 80 kHz is demonstrated with both grazing and normal-incidence geometries. The substitution-based method, as opposed to a simultaneous method, avoids problems associated with the nonuniformity of the sound field and, as applied here, uses a 1/4-in. air-condenser pressure microphone as a known reference. Best results were obtained with a centrifugal fan, which is used as a random, broadband sound source. A broadband source minimizes reflection-related interferences that can plague FF measurements. Calibrations were performed on 1/4-in. FF air-condenser, electret, and microelectromechanical systems (MEMS) microphones in an anechoic chamber. The uncertainty of this FF method is estimated by comparing the pressure sensitivity of an air-condenser FF microphone, as derived from the FF measurement, with that of an electrostatic actuator calibration. The root-mean-square difference is found to be +/- 0.3 dB over the range 1-80 kHz, and the combined standard uncertainty of the FF method, including other significant contributions, is +/- 0.41 dB. PMID:16454287

  1. CBaM : A calibration, bridging and merging method for post-processing GCM forecasts of meteorological variables

    NASA Astrophysics Data System (ADS)

    Schepen, A.; Wang, Q. J.; Robertson, D. E.

    2014-12-01

    A calibration, bridging and merging (CBaM) method has been developed to post-process outputs from general circulation models (GCMs) for seasonal forecasting of climate variables. An overview of the methodology and a summary of applications will be given. Post-processing of GCM forecasts is often necessary for the outputs to be more informative. CBaM attempts to maximise the value of GCM outputs by not only post-processing the variable of interest (calibration), but also using other available outputs such as sea surface temperatures to generate forecasts (bridging). Merging forecasts from calibration and bridging models leads to the opportunity to improve forecasting skill for some regions and time periods. In CBaM, separate calibration and bridging models are established using a Bayesian joint probability modelling approach. The models generate forecasts in the form of ensembles. Forecasts from multiple calibration and bridging models are merged using Bayesian model averaging. Ensemble time series forecasts are produced by sequencing ensemble members using the Schaake Shuffle. Results to date are presented for a number of applications. The method is applied to produce gridded rainfall forecasts for Australia and China, using outputs from single or multiple GCMs. It is also applied to produce monthly forecasts of catchment rainfall for up to 12 months in advance. Monthly forecasts of catchment rainfall are used in a hydrological model to forecast streamflow for up to 12 months. CBaM forecasts are shown to extract skill from the atmospheric and oceanic modules of the GCM, and are also shown to be reliable. Work is in progress to apply CBaM to forecasts of other climate variables, including temperature, and to combine forecasts from multiple GCMs, including the ECMWF System 4 and NCEP CFSv2 models.

  2. Viking lander camera geometry calibration report. Volume 1: Test methods and data reduction techniques

    NASA Technical Reports Server (NTRS)

    Wolf, M. B.

    1981-01-01

    The determination and removal of instrument signature from Viking Lander camera geometric data are described. All tests conducted as well as a listing of the final database (calibration constants) used to remove instrument signature from Viking Lander flight images are included. The theory of the geometric aberrations inherent in the Viking Lander camera is explored.

  3. Method and apparatus of a portable imaging-based measurement with self calibration

    DOEpatents

    Chang, Tzyy-Shuh; Huang, Hsun-Hau

    2012-07-31

    A portable imaging-based measurement device is developed to perform 2D projection based measurements on an object that is difficult or dangerous to access. This device is equipped with self calibration capability and built-in operating procedures to ensure proper imaging based measurement.

  4. A novel plane method to the calibration of the thermal camera

    NASA Astrophysics Data System (ADS)

    Wang, Xunsi; Huang, Wei; Nie, Qiu-hua; Xu, Tiefeng; Dai, Shixun; Shen, Xiang; Cheng, Weihai

    2009-07-01

    This paper provides an up-to-date review of research efforts in thermal camera and target object recognition techniques based on two-dimensional (2D) images in the infrared (IR) spectra (8-12μm). From the geometric point of view, a special target plate was constructed with a radiation source of lamp excited that allows all of these devices to be calibrated geometrically along a radiance-based approach. The calibration theory and actual experimental procedures were described, then an automated measurement of the circle targets by image centroid algorithm. The key parameters of IR camera were calibrated out with 3 inner and 6 outer of Tsai model in thermal imaging. The subsequent data processing and analysis were then outlined. The 3D model from the successful calibration of a representative sample of the infrared array camera was presented and discussed. They provide much new and easy way to the geometric characteristics of these imagers that can be used in car-night-vision, medical, industrial, military, and environmental applications.

  5. Apparatus for measuring resistance change only in a cell analyzer and method for calibrating it

    DOEpatents

    Hoffman, Robert A.

    1980-01-01

    The disclosure relates to resistance only monitoring and calibration in an electrical cell analyzer. Sample and sheath fluid flows of different salinities are utilized, the sample flow being diameter modulated to produce a selected pattern which is compared to the resistance measured across the flows.

  6. Application of the Langley plot method to the calibration of the solar backscattered ultraviolet instrument on the Nimbus 7 satellite

    NASA Technical Reports Server (NTRS)

    Bhartia, P. K.; Taylor, S.; Mcpeters, R. D.; Wellemeyer, C.

    1995-01-01

    The concept of the well-known Langley plot technique, used for the calibration of ground-based instruments, has been generalized for application to satellite instruments. In polar regions, near summer solstice, the solar backscattered ultraviolet (SBUV) instrument on the Nimbus 7 satellite samples the same ozone field at widely different solar zenith angles. These measurements are compared to assess the long-term drift in the instrument calibration. Although the technique provides only a relative wavelength-to-wavelength calibration, it can be combined with existing techniques to determine the drift of the instrument at any wavelength. Using this technique, we have generated a 12-year data set of ozone vertical profiles from SBUV with an estimated accuracy of +/- 5% at 1 mbar and +/- 2% at 10 mbar (95% confidence) over 12 years. Since the method is insensitive to true changes in the atmospheric ozone profile, it can also be used to compare the calibrations of similar SBUV instruments launched without temporal overlap.

  7. TU-A-18A-01: Basic Principles of PET/CT, Calibration Methods and Contrast Recovery Across Multiple Cameras

    SciTech Connect

    Kappadath, S; Nye, J

    2014-06-15

    This continuing education session will discuss the physical principles of PET/CT imaging and characterization of contrast recovery using accreditation phantoms. A detailed overview will be given on the physical principles of PET including positron decay physics, 2D and 3D data acquisition, time-of-flight, scatter correction, CT attenuation correction, and image reconstruction. Instrument quality control and calibration procedures will be discussed. Technical challenges, common image artifacts and strategies to mitigate these issues will also be discussed. Data will be presented on acquisition techniques and reconstruction parameters affecting contrast recovery. The discussion will emphasize the minimization of reconstruction differences in quantification metrics such as SUV and contrast recovery coefficients for the NEMA and ACR clinical trial phantoms. Data from new and older generation scanners will be shown including comparison of contrast recovery measurements to their analytical solutions. The goal of this session is to update attendees on the quality control and calibration of PET/CT scanners, on methods to establish a common calibration for PET/CT scanners to control for instrument variance across multiple sites. Learning Objectives: Review the physical principles of PET/CT, quality control and calibration Gain further understanding on how to apply techniques for improving quantitative agreement across multiple cameras Describe the differences between measured and expected contrast recovery for the NEMA and ACR PET phantoms.

  8. Comparison of official methods for 'readily oxidizable substances' in propionic acid as a food additive.

    PubMed

    Ishiwata, H; Takeda, Y; Kawasaki, Y; Kubota, H; Yamada, T

    1996-01-01

    The official methods for 'readily oxidizable substances (ROS)' in propionic acid as a food additive were compared. The methods examined were those adopted in the Compendium of Food Additive Specifications (CFAS) by the Joint FAO-WHO Expert Committee on Food Additives, FAO, The Japanese Standards for Food Additives (JSFA) by the Ministry of Health and Welfare, Japan, and the Food Chemicals Codex (FCC) by the National Research Council, USA. The methods given in CFAS and JSFA are the same (potassium permanganate consumption). However, by this method, manganese (VII) in potassium permanganate was readily reduced to colourless manganese(II) with some substances contained in the propionic acid before reacting with aldehydes, which are generally considered as 'readily oxidizable substances', to form brown manganese (IV) oxide. The FCC method (bromine consumption) for 'ROS' could be recommended because it was able to obtain quantitative results of 'ROS', including aldehydes. PMID:8647299

  9. Calculation of depth-dependent elemental concentration with X-ray fluorescence using a layered calibration method

    NASA Astrophysics Data System (ADS)

    Gherase, M. R.; Fleming, D. E. B.

    2011-05-01

    Depth-dependent elemental concentration can be assessed using either multiple-angle or confocal X-ray fluorescence (XRF) measurements. This work presents a different approach based on a layered calibration method. The depth-dependent elemental concentration was modeled as multiple layers of uniform elemental concentrations. The unknown elemental concentration in each layer can be calculated using a single-angle XRF measurement, layered calibration data, and a priori knowledge of the concentration behavior as a function of depth. The method was verified using a commercial portable X-ray spectrometer and four-layer stacks of polyester resin discs doped with various concentrations of arsenic. This approach is particularly suitable for quantitative in vivo measurements of arsenic and selenium concentrations in the human skin since minimal ionizing radiation exposure constrains the number of XRF measurements.

  10. Standard addition method for the determination of pharmaceutical residues in drinking water by SPE-LC-MS/MS.

    PubMed

    Cimetiere, Nicolas; Soutrel, Isabelle; Lemasle, Marguerite; Laplanche, Alain; Crocq, André

    2013-01-01

    The study of the occurrence and fate of pharmaceutical compounds in drinking or waste water processes has become very popular in recent years. Liquid chromatography with tandem mass spectrometry is a powerful analytical tool often used to determine pharmaceutical residues at trace level in water. However, many steps may disrupt the analytical procedure and bias the results. A list of 27 environmentally relevant molecules, including various therapeutic classes and (cardiovascular, veterinary and human antibiotics, neuroleptics, non-steroidal anti-inflammatory drugs, hormones and other miscellaneous pharmaceutical compounds), was selected. In this work, a method was developed using ultra performance liquid chromatography coupled to tandem mass spectrometry (UPLC-MS/MS) and solid-phase extraction to determine the concentration of the 27 targeted pharmaceutical compounds at the nanogram per litre level. The matrix effect was evaluated from water sampled at different treatment stages. Conventional methods with external calibration and internal standard correction were compared with the standard addition method (SAM). An accurate determination of pharmaceutical compounds in drinking water was obtained by the SAM associated with UPLC-MS/MS. The developed method was used to evaluate the occurrence and fate of pharmaceutical compounds in some drinking water treatment plants in the west of France. PMID:24617062

  11. Oil palm water use: calibration of a sap flux method and a field measurement scheme.

    PubMed

    Niu, Furong; Röll, Alexander; Hardanto, Afik; Meijide, Ana; Köhler, Michael; Hendrayanto; Hölscher, Dirk

    2015-05-01

    Oil palm (Elaeis guineensis Jacq.) water use was assessed by sap flux density measurements with the aim to establish the method and derive water-use characteristics. Thermal dissipation probes were inserted into leaf petioles of mature oil palms. In the laboratory, we tested our set-up against gravimetric measurements and derived new parameters for the original calibration equation that are specific to oil palm petioles. In the lowlands of Jambi, Indonesia, in a 12-year-old monoculture plantation, 56 leaves on 10 palms were equipped with one sensor per leaf. A 10-fold variation in individual leaf water use among leaves was observed, but we did not find significant correlations to the variables trunk height and diameter, leaf azimuthal orientation, leaf inclination or estimated horizontal leaf shading. We thus took an un-stratified approach to determine an appropriate sampling design to estimate stand transpiration (Es, mm day(-1)) rates of oil palm. We used the relative standard error of the mean (SEn, %) as a measure for the potential estimation error of Es associated with sample size. It was 14% for a sample size of 13 leaves to determine the average leaf water use and four palms to determine the average number of leaves per palm. Increasing these sample sizes only led to minor further decreases of the SEn of Es. The observed 90-day average of Es was 1.1 mm day(-1) (error margin ± 0.2 mm day(-1)), which seems relatively low, but does not contradict Penman-Monteith-derived estimates of evapotranspiration. Examining the environmental drivers of Es on an intra-daily scale indicates an early, pre-noon maximum of Es rates (11 am) due to a very sensitive reaction of Es to increasing vapor pressure deficit in the morning. This early peak is followed by a steady decline of Es rates for the rest of the day, despite further rising levels of vapor pressure deficit and radiation; this results in pronounced hysteresis, particularly between Es and vapor pressure deficit. PMID

  12. Biological chromodynamics: a general method for measuring protein occupancy across the genome by calibrating ChIP-seq

    PubMed Central

    Hu, Bin; Petela, Naomi; Kurze, Alexander; Chan, Kok-Lung; Chapard, Christophe; Nasmyth, Kim

    2015-01-01

    Sequencing DNA fragments associated with proteins following in vivo cross-linking with formaldehyde (known as ChIP-seq) has been used extensively to describe the distribution of proteins across genomes. It is not widely appreciated that this method merely estimates a protein's distribution and cannot reveal changes in occupancy between samples. To do this, we tagged with the same epitope orthologous proteins in Saccharomyces cerevisiae and Candida glabrata, whose sequences have diverged to a degree that most DNA fragments longer than 50 bp are unique to just one species. By mixing defined numbers of C. glabrata cells (the calibration genome) with S. cerevisiae samples (the experimental genomes) prior to chromatin fragmentation and immunoprecipitation, it is possible to derive a quantitative measure of occupancy (the occupancy ratio – OR) that enables a comparison of occupancies not only within but also between genomes. We demonstrate for the first time that this ‘internal standard’ calibration method satisfies the sine qua non for quantifying ChIP-seq profiles, namely linearity over a wide range. Crucially, by employing functional tagged proteins, our calibration process describes a method that distinguishes genuine association within ChIP-seq profiles from background noise. Our method is applicable to any protein, not merely highly conserved ones, and obviates the need for the time consuming, expensive, and technically demanding quantification of ChIP using qPCR, which can only be performed on individual loci. As we demonstrate for the first time in this paper, calibrated ChIP-seq represents a major step towards documenting the quantitative distributions of proteins along chromosomes in different cell states, which we term biological chromodynamics. PMID:26130708

  13. Biological chromodynamics: a general method for measuring protein occupancy across the genome by calibrating ChIP-seq.

    PubMed

    Hu, Bin; Petela, Naomi; Kurze, Alexander; Chan, Kok-Lung; Chapard, Christophe; Nasmyth, Kim

    2015-11-16

    Sequencing DNA fragments associated with proteins following in vivo cross-linking with formaldehyde (known as ChIP-seq) has been used extensively to describe the distribution of proteins across genomes. It is not widely appreciated that this method merely estimates a protein's distribution and cannot reveal changes in occupancy between samples. To do this, we tagged with the same epitope orthologous proteins in Saccharomyces cerevisiae and Candida glabrata, whose sequences have diverged to a degree that most DNA fragments longer than 50 bp are unique to just one species. By mixing defined numbers of C. glabrata cells (the calibration genome) with S. cerevisiae samples (the experimental genomes) prior to chromatin fragmentation and immunoprecipitation, it is possible to derive a quantitative measure of occupancy (the occupancy ratio - OR) that enables a comparison of occupancies not only within but also between genomes. We demonstrate for the first time that this 'internal standard' calibration method satisfies the sine qua non for quantifying ChIP-seq profiles, namely linearity over a wide range. Crucially, by employing functional tagged proteins, our calibration process describes a method that distinguishes genuine association within ChIP-seq profiles from background noise. Our method is applicable to any protein, not merely highly conserved ones, and obviates the need for the time consuming, expensive, and technically demanding quantification of ChIP using qPCR, which can only be performed on individual loci. As we demonstrate for the first time in this paper, calibrated ChIP-seq represents a major step towards documenting the quantitative distributions of proteins along chromosomes in different cell states, which we term biological chromodynamics. PMID:26130708

  14. Measurements of Temperature Variations in the Atmosphere near the Tropopause with Reference to Airspeed Calibration by the Temperature Method

    NASA Technical Reports Server (NTRS)

    Lina, Lindsay J; Ricker, Harry H , Jr

    1952-01-01

    Measurements of temperature variations in the atmosphere near the tropopause over land in the vicinity of Langley Field, Va., are presented. This investigation was made for the purpose of obtaining information on the accuracy of the temperature method(NACA TN 2046) of airspeed calibration over the range of Mach number from 0.6 to .08. The temperature surveys and the description of a specially designed thermometer are also presented.

  15. New calibration method using low cost MEM IMUs to verify the performance of UAV-borne MMS payloads.

    PubMed

    Chiang, Kai-Wei; Tsai, Meng-Lun; Naser, El-Sheimy; Habib, Ayman; Chu, Chien-Hsun

    2015-01-01

    Spatial information plays a critical role in remote sensing and mapping applications such as environment surveying and disaster monitoring. An Unmanned Aerial Vehicle (UAV)-borne mobile mapping system (MMS) can accomplish rapid spatial information acquisition under limited sky conditions with better mobility and flexibility than other means. This study proposes a long endurance Direct Geo-referencing (DG)-based fixed-wing UAV photogrammetric platform and two DG modules that each use different commercial Micro-Electro Mechanical Systems' (MEMS) tactical grade Inertial Measurement Units (IMUs). Furthermore, this study develops a novel kinematic calibration method which includes lever arms, boresight angles and camera shutter delay to improve positioning accuracy. The new calibration method is then compared with the traditional calibration approach. The results show that the accuracy of the DG can be significantly improved by flying at a lower altitude using the new higher specification hardware. The new proposed method improves the accuracy of DG by about 20%. The preliminary results show that two-dimensional (2D) horizontal DG positioning accuracy is around 5.8 m at a flight height of 300 m using the newly designed tactical grade integrated Positioning and Orientation System (POS). The positioning accuracy in three-dimensions (3D) is less than 8 m. PMID:25808764

  16. A New Calibration Method Using Low Cost MEM IMUs to Verify the Performance of UAV-Borne MMS Payloads

    PubMed Central

    Chiang, Kai-Wei; Tsai, Meng-Lun; Naser, El-Sheimy; Habib, Ayman; Chu, Chien-Hsun

    2015-01-01

    Spatial information plays a critical role in remote sensing and mapping applications such as environment surveying and disaster monitoring. An Unmanned Aerial Vehicle (UAV)-borne mobile mapping system (MMS) can accomplish rapid spatial information acquisition under limited sky conditions with better mobility and flexibility than other means. This study proposes a long endurance Direct Geo-referencing (DG)-based fixed-wing UAV photogrammetric platform and two DG modules that each use different commercial Micro-Electro Mechanical Systems’ (MEMS) tactical grade Inertial Measurement Units (IMUs). Furthermore, this study develops a novel kinematic calibration method which includes lever arms, boresight angles and camera shutter delay to improve positioning accuracy. The new calibration method is then compared with the traditional calibration approach. The results show that the accuracy of the DG can be significantly improved by flying at a lower altitude using the new higher specification hardware. The new proposed method improves the accuracy of DG by about 20%. The preliminary results show that two-dimensional (2D) horizontal DG positioning accuracy is around 5.8 m at a flight height of 300 m using the newly designed tactical grade integrated Positioning and Orientation System (POS). The positioning accuracy in three-dimensions (3D) is less than 8 m. PMID:25808764

  17. Gamma spectrometry efficiency calibration using Monte Carlo methods to measure radioactivity of 137Cs in food samples.

    PubMed

    Alrefae, T

    2014-12-01

    A simple method of efficiency calibration for gamma spectrometry was performed. This method, which focused on measuring the radioactivity of (137)Cs in food samples, was based on Monte Carlo simulations available in the free-of-charge toolkit GEANT4. Experimentally, the efficiency values of a high-purity germanium detector were calculated for three reference materials representing three different food items. These efficiency values were compared with their counterparts produced by a computer code that simulated experimental conditions. Interestingly, the output of the simulation code was in acceptable agreement with the experimental findings, thus validating the proposed method. PMID:24214912

  18. White-light Interferometry using a Channeled Spectrum: II. Calibration Methods, Numerical and Experimental Results

    NASA Technical Reports Server (NTRS)

    Zhai, Chengxing; Milman, Mark H.; Regehr, Martin W.; Best, Paul K.

    2007-01-01

    In the companion paper, [Appl. Opt. 46, 5853 (2007)] a highly accurate white light interference model was developed from just a few key parameters characterized in terms of various moments of the source and instrument transmission function. We develop and implement the end-to-end process of calibrating these moment parameters together with the differential dispersion of the instrument and applying them to the algorithms developed in the companion paper. The calibration procedure developed herein is based on first obtaining the standard monochromatic parameters at the pixel level: wavenumber, phase, intensity, and visibility parameters via a nonlinear least-squares procedure that exploits the structure of the model. The pixel level parameters are then combined to obtain the required 'global' moment and dispersion parameters. The process is applied to both simulated scenarios of astrometric observations and to data from the microarcsecond metrology testbed (MAM), an interferometer testbed that has played a prominent role in the development of this technology.

  19. Probabilistic methods for sensitivity analysis and calibration in the NASA challenge problem

    SciTech Connect

    Safta, Cosmin; Sargsyan, Khachik; Najm, Habib N.; Chowdhary, Kenny; Debusschere, Bert; Swiler, Laura P.; Eldred, Michael S.

    2015-01-01

    In this study, a series of algorithms are proposed to address the problems in the NASA Langley Research Center Multidisciplinary Uncertainty Quantification Challenge. A Bayesian approach is employed to characterize and calibrate the epistemic parameters based on the available data, whereas a variance-based global sensitivity analysis is used to rank the epistemic and aleatory model parameters. A nested sampling of the aleatory–epistemic space is proposed to propagate uncertainties from model parameters to output quantities of interest.

  20. Simplified stereo-optical ultrasound plane calibration

    NASA Astrophysics Data System (ADS)

    Hoßbach, Martin; Noll, Matthias; Wesarg, Stefan

    2013-03-01

    Image guided therapy is a natural concept and commonly used in medicine. In anesthesia, a common task is the injection of an anesthetic close to a nerve under freehand ultrasound guidance. Several guidance systems exist using electromagnetic tracking of the ultrasound probe as well as the needle, providing the physician with a precise projection of the needle into the ultrasound image. This, however, requires additional expensive devices. We suggest using optical tracking with miniature cameras attached to a 2D ultrasound probe to achieve a higher acceptance among physicians. The purpose of this paper is to present an intuitive method to calibrate freehand ultrasound needle guidance systems employing a rigid stereo camera system. State of the art methods are based on a complex series of error prone coordinate system transformations which makes them susceptible to error accumulation. By reducing the amount of calibration steps to a single calibration procedure we provide a calibration method that is equivalent, yet not prone to error accumulation. It requires a linear calibration object and is validated on three datasets utilizing di erent calibration objects: a 6mm metal bar and a 1:25mm biopsy needle were used for experiments. Compared to existing calibration methods for freehand ultrasound needle guidance systems, we are able to achieve higher accuracy results while additionally reducing the overall calibration complexity. Ke