Sample records for fluoroscopy procedures realized

  1. Feasibility Study of Needle Placement in Percutaneous Vertebroplasty: Cone-Beam Computed Tomography Guidance Versus Conventional Fluoroscopy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Braak, Sicco J., E-mail: sjbraak@gmail.com; Zuurmond, Kirsten, E-mail: kirsten.zuurmond@philips.com; Aerts, Hans C. J., E-mail: hans.cj.aerts@philips.com

    2013-08-01

    ObjectiveTo investigate the accuracy, procedure time, fluoroscopy time, and dose area product (DAP) of needle placement during percutaneous vertebroplasty (PVP) using cone-beam computed tomography (CBCT) guidance versus fluoroscopy.Materials and MethodsOn 4 spine phantoms with 11 vertebrae (Th7-L5), 4 interventional radiologists (2 experienced with CBCT guidance and two inexperienced) punctured all vertebrae in a bipedicular fashion. Each side was randomization to either CBCT guidance or fluoroscopy. CBCT guidance is a sophisticated needle guidance technique using CBCT, navigation software, and real-time fluoroscopy. The placement of the needle had to be to a specific target point. After the procedure, CBCT was performed tomore » determine the accuracy, procedure time, fluoroscopy time, and DAP. Analysis of the difference between methods and experience level was performed.ResultsMean accuracy using CBCT guidance (2.61 mm) was significantly better compared with fluoroscopy (5.86 mm) (p < 0.0001). Procedure time was in favor of fluoroscopy (7.39 vs. 10.13 min; p = 0.001). Fluoroscopy time during CBCT guidance was lower, but this difference is not significant (71.3 vs. 95.8 s; p = 0.056). DAP values for CBCT guidance and fluoroscopy were 514 and 174 mGy cm{sup 2}, respectively (p < 0.0001). There was a significant difference in favor of experienced CBCT guidance users regarding accuracy for both methods, procedure time of CBCT guidance, and added DAP values for fluoroscopy.ConclusionCBCT guidance allows users to perform PVP more accurately at the cost of higher patient dose and longer procedure time. Because procedural complications (e.g., cement leakage) are related to the accuracy of the needle placement, improvements in accuracy are clinically relevant. Training in CBCT guidance is essential to achieve greater accuracy and decrease procedure time/dose values.« less

  2. Zero-fluoroscopy permanent pacemaker implantation using Ensite NavX system: Clinical viability or fanciful technique?

    PubMed

    Guo, Ping; Qiu, Jie; Wang, Yan; Chen, Guangzhi; Proietti, Riccardo; Fadhle, Al-Selmi; Zhao, Chunxia; Wen Wang, Dao

    2018-02-01

    Fluoroscopy is the imaging modality routinely used for cardiac device implantation and electrophysiological procedures. Due to the rising concern regarding the harmful effects of radiation exposure to both the patients and operation staffs, novel 3D mapping systems have been developed and implemented in electrophysiological procedure for the navigation of catheters inside the heart chambers. Their applicability in cardiac device implantation has been rarely reported. Our aim is to evaluate the feasibility and safety of permanent pacemaker implantation without fluoroscopy. From January 2012 to June 2016, six patients (50 ± 15 years, four of six were female, one of who was at the 25th week of gestation) who underwent permanent pacemaker implantation were included (zero-fluoroscopy group). Data from 20 consecutive cases of implantation performed under fluoroscopy guidance were chosen as a control group (fluoroscopy group). Total implantation procedure time for single-chamber pacemaker was 51.3 ± 13.1 minutes in the zero-fluoroscopy group and 42.6 ± 7.4 minutes in the fluoroscopy group (P  =  0.155). The implantation procedural time for a dual-chamber pacemaker was 88.3 ± 19.6 minutes and 67.3 ± 7.6 minutes in the zero-fluoroscopy and fluoroscopy groups (P  =  0.013), respectively. No complications were observed during the procedure and the follow-up in the two groups, and all pacemakers worked with satisfactory parameters. Ensite NavX system can be used as a reliable and safe zero-fluoroscopy approach for the implantation of single- or dual-chamber permanent pacemakers in specific patients, such as pregnant women or in extreme situations when the x-ray machine is not available. © 2017 The Authors. Pacing and Clinical Electrophysiology published by Wiley Periodicals, Inc.

  3. Scattered radiation risk to the lens of the eyes for staff involved in using mobile C-arm fluoroscopy unit: Which position is riskiest?

    NASA Astrophysics Data System (ADS)

    Salleh, H.; Samat, S. B.; Matori, M. K.; Isa, M. J. M.

    2015-09-01

    Cataractogenesis is something to be concerned by radiologist and radiographer who work extensively in fluoroscopy. The increasing use of fluoroscopy or interventional fluoroscopy has to come with safety awareness on scattered radiation risk for staff performing the procedure. This study is looking into the radiation risk to the lens of the eyes for staff involved in fluoroscopy using the mobile C-arm fluoroscopy unit. The Toshiba SXT-1000A and Alderson Rando phantom were used in this study. Based on the results, it is found clearly that over couch (OC) procedure is riskier than under couch (UC) procedure. The cathode bound area is clearly riskier than anode bound area especially for UC procedure. More doses (at least +1,568 % of safest position) are received by the lens of the eyes for staff standing at the cathode bound area especially the position opposite to the x-ray tube.

  4. Scattered radiation risk to the lens of the eyes for staff involved in using mobile C-arm fluoroscopy unit: Which position is riskiest?

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Salleh, H.; Matori, M. K.; Isa, M. J. M.

    Cataractogenesis is something to be concerned by radiologist and radiographer who work extensively in fluoroscopy. The increasing use of fluoroscopy or interventional fluoroscopy has to come with safety awareness on scattered radiation risk for staff performing the procedure. This study is looking into the radiation risk to the lens of the eyes for staff involved in fluoroscopy using the mobile C-arm fluoroscopy unit. The Toshiba SXT-1000A and Alderson Rando phantom were used in this study. Based on the results, it is found clearly that over couch (OC) procedure is riskier than under couch (UC) procedure. The cathode bound area ismore » clearly riskier than anode bound area especially for UC procedure. More doses (at least +1,568 % of safest position) are received by the lens of the eyes for staff standing at the cathode bound area especially the position opposite to the x-ray tube.« less

  5. Feasibility of zero or near zero fluoroscopy during catheter ablation procedures.

    PubMed

    Haegeli, Laurent M; Stutz, Linda; Mohsen, Mohammed; Wolber, Thomas; Brunckhorst, Corinna; On, Chol-Jun; Duru, Firat

    2018-04-03

    Awareness of risks associated with radiation exposure to patients and medical staff has significantly increased. It has been reported before that the use of advanced three-dimensional electro-anatomical mapping (EAM) system significantly reduces fluoroscopy time, however this study aimed for zero or near zero fluoroscopy ablation to assess its feasibility and safety in ablation of atrial fibrillation (AF) and other tachyarrhythmias in a "real world" experience of a single tertiary care center. This was a single-center study where ablation procedures were attempted without fluoroscopy in 34 consecutive patients with different tachyarrhythmias under the support of EAM system. When transseptal puncture (TSP) was needed, it was attempted under the guidance of intracardiac echocardiography (ICE). Among 34 patients consecutively enrolled in this study, 28 (82.4%) patients were referred for radiofrequency ablation (RFA) of AF, 3 (8.8%) patients for ablation of right ventricular outflow tract (RVOT) ventricular extrasystole (VES), 1 (2.9%) patient for ablation of atrioventricular nodal reentry tachycardia (AVNRT), 2 (5.9%) patients for typical atrial flutter ablation. In 21 (62%) patients the entire procedure was carried out without the use of fluoroscopy. Among 28 AF patients, 15 (54%) patients underwent ablation without the use of fluoroscopy and among these 15 patients, 10 (67%) patients required TSP under ICE guidance while 5 (33%) patients the catheters were introduced to left atrium through a patent foramen ovale. In 13 AF patients, fluoroscopy was only required for double TSP. The total procedure time of AF ablation was 130 ± 50 min. All patients referred for atrial flutter, AVNRT, and VES of the RVOT ablation did not require any fluoroscopy. This study demonstrates the feasibility of zero or near zero fluoroscopy procedure including TSP with the support of EAM and ICE guidance in a "real world" experience of a single tertiary care center. When fluoroscopy was required, it was limited to TSP hence keeping the radiation dose very low. .

  6. Radiation exposure, and procedure and fluoroscopy times in endovascular treatment of intracranial aneurysms: a methodological comparison.

    PubMed

    Cheung, Nicholas K; Boutchard, Michelle; Carr, Michael W; Froelich, Jens J

    2018-01-09

    Limited data are available for radiation exposure, and procedure and fluoroscopy times in neuroendovascular treatment (NET) strategies. This study establishes and compares related parameters between coil embolization (COIL), balloon assisted coil embolization (BAC), stent assisted coil embolization (SAC), and flow diverting technology (FDT) in NET of intracranial aneurysms. Between 2010 and 2017, 249 consecutive intracranial aneurysms underwent NET at a single center, all performed by the same operator. Dose area products (DAP), and procedure and fluoroscopy times were recorded and compared between COIL, BAC, SAC, and FDT techniques. Differences in parameters between cohorts were analyzed for significance using the Mann-Whitney U test, unpaired t test and χ 2 test. Additional subgroup analysis was performed for emergency and elective cases. 83 aneurysms were treated with COIL (33%), 72 with BAC (29%), 61 with SAC (25%), and 33 with FDT (13%). Baseline characteristics were largely similar within these groups (P>0.05). Among COIL, BAC, and FDT cohorts, no significant difference was found for mean DAP, or procedure and fluoroscopy times (P>0.05). However, compared with all other cohorts, SAC was associated with a significantly higher DAP and longer procedure and fluoroscopy times (P<0.005). No significant difference was recorded for emergency and elective case subgroups. Compared with other NET strategies, SAC was associated with a significantly higher DAP, and longer procedure and fluoroscopy times. This study provides an initial dataset regarding radiation exposure, and procedure and fluoroscopy times for common NET, and may assist ALARA (As Low As Reasonably Achievable) principles to reduce radiation risks. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2018. All rights reserved. No commercial use is permitted unless otherwise expressly granted.

  7. Digital methods for reducing radiation exposure during medical fluoroscopy

    NASA Astrophysics Data System (ADS)

    Edmonds, Ernest W.; Rowlands, John A.; Hynes, David M.; Toth, B. D.; Porter, Anthony J.

    1990-07-01

    There is increased concern over radiation exposure to the general population from many sources. One of the most significant sources is that received by the patient during medical diagnostic procedures, and of these, the procedure with the greatest potential hazard is fluoroscopy. The legal limit for fluoroscopy in most jurisdictions is SR per minute skin exposure rate. Fluoroscopes are often operated in excess of this figure, and in the case of interventional procedures, fluorocopy times may exceed 20 minutes. With improvements in medical technology these procedures are being performed more often, and also are being carried out on younger age groups. Radiation exposure during fluoroscopy, both to patient and operator, is therefore becoming a matter of increasing concern to regulating authorities, and it is incumbent on us to develop digital technology to minimise the radiation hazard in these procedures. This paper explores the technical options available for radiation exposure reduction, including pulsed fluoroscopy, digital noise reduction, or simple reduction in exposure rate to the x-ray image intensifier. We also discuss educational aspects of fluoroscopy which radiologists should be aware of which can be more important than the technological solutions. A "work in progress" report gives a completely new approach to the implementation of a large number of possible digital algorithms, for the investigation of clinical efficacy.

  8. Radiation exposure from fluoroscopy during orthopedic surgical procedures

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Riley, S.A.

    1989-11-01

    The use of fluoroscopy has enabled orthopedic surgeons to become technically more proficient. In addition, these surgical procedures tend to have less associated patient morbidity by decreasing operative time and minimizing the area of the operative field. The trade-off, however, may be an increased risk of radiation exposure to the surgeon on an annual or lifetime basis. The current study was designed to determine the amount of radiation received by the primary surgeon and the first assistant during selected surgical procedures involving the use of fluoroscopy. Five body sites exposed to radiation were monitored for dosage. The results of thismore » study indicate that with appropriate usage, (1) radiation exposure from fluoroscopy is relatively low; (2) the surgeon's dominant hand receives the most exposure per case; and (3) proper maintenance and calibration of fluoroscopic machines are important factors in reducing exposure risks. Therefore, with proper precautions, the use of fluoroscopy in orthopedic procedures can remain a safe practice.« less

  9. Impact of attributed audit on procedural performance in cardiac electrophysiology catheter laboratory.

    PubMed

    Sawhney, V; Volkova, E; Shaukat, M; Khan, F; Segal, O; Ahsan, S; Chow, A; Ezzat, V; Finlay, M; Lambiase, P; Lowe, M; Dhinoja, M; Sporton, S; Earley, M J; Hunter, R J; Schilling, R J

    2018-06-01

    Audit has played a key role in monitoring and improving clinical practice. However, audit often fails to drive change as summative institutional data alone may be insufficient to do so. We hypothesised that the practice of attributed audit, wherein each individual's procedural performance is presented will have a greater impact on clinical practice. This hypothesis was tested in an observational study evaluating improvement in fluoroscopy times for AF ablation. Retrospective analyses of fluoroscopy times in AF ablations at the Barts Heart Centre (BHC) from 2012-2017. Fluoroscopy times were compared pre- and post- the introduction of attributed audit in 2012 at St Bartholomew's Hospital (SBH). In order to test the hypothesis, this concept was introduced to a second group of experienced operators from the Heart Hospital (HH) as part of a merger of the two institutions in 2015 and change in fluoroscopy times recorded. A significant drop in fluoroscopy times (33.3 ± 9.14 to 8.95 ± 2.50, p < 0.0001) from 2012-2014 was noted after the introduction of attributed audit. At the time of merger, a significant difference in fluoroscopy times between operators from the two centres was seen in 2015. Each operator's procedural performance was shared openly at the audit meeting. Subsequent audits showed a steady decrease in fluoroscopy times for each operator with the fluoroscopy time (min, mean±SD) decreasing from 13.29 ± 7.3 in 2015 to 8.84 ± 4.8 (p < 0.0001) in 2017 across the entire group. Systematic improvement in fluoroscopy times for AF ablation procedures was noted byevaluating individual operators' performance. Attributing data to physicians in attributed audit can promptsignificant improvement and hence should be adopted in clinical practice.

  10. Implementation of a near-zero fluoroscopy approach in interventional electrophysiology: impact of operator experience.

    PubMed

    Wannagat, Severin; Loehr, Lena; Lask, Sebastian; Völk, Katharina; Karaköse, Tamer; Özcelik, Cemil; Mügge, Andreas; Wutzler, Alexander

    2018-04-01

    Catheter ablation is performed under fluoroscopic guidance. Reduction of radiation dose for patients and staff is emphasized by current recommendations. Previous studies have shown that lower operator experience leads to increased radiation dose. On the other hand, less experienced operators may depend even more on fluoroscopic guidance. Our study aimed to evaluate feasibility and efficacy of a non-fluoroscopic approach in different training levels. From January 2017, a near-zero fluoroscopy approach was established in two centers. Four operators (beginner, 1st year fellow, 2nd year fellow, expert) were instructed to perform the complete procedure with the use of a 3-D mapping system without fluoroscopy. A historical cohort that underwent procedures with fluoroscopy use served as control group. Dose area product (DPA), procedure duration, acute procedural success, and complications were compared between the groups and for each operator. Procedures were performed in 157 patients. The first 100 patients underwent procedures with fluoroscopic guidance, the following 57 procedures were performed with the near-zero fluoroscopy approach. The results show a significant reduction in DPA for all operators immediately after implementation of the near-zero fluoroscopy protocol (control 637 ± 611 μGy/m 2 ; beginner 44.1 ± 79.5 μGy/m 2 , p = 0.002; 1st year fellow 24.3 ± 46.4.5 μGy/m 2 , p = 0.001; 2nd year fellow 130.3 ± 233.3 μGy/m 2 , p = 0.003; expert 9.3 ± 37.4 μGy/m 2 , P < 0.001). Procedure duration, acute success, and complications were not significantly different between the groups. Our results show a 90% reduction of DPA shortly after implementation of a near-zero fluoroscopy approach in interventional electrophysiology even in operators in training.

  11. Estimating effective dose to pediatric patients undergoing interventional radiology procedures using anthropomorphic phantoms and MOSFET dosimeters.

    PubMed

    Miksys, Nelson; Gordon, Christopher L; Thomas, Karen; Connolly, Bairbre L

    2010-05-01

    The purpose of this study was to estimate the effective doses received by pediatric patients during interventional radiology procedures and to present those doses in "look-up tables" standardized according to minute of fluoroscopy and frame of digital subtraction angiography (DSA). Organ doses were measured with metal oxide semiconductor field effect transistor (MOSFET) dosimeters inserted within three anthropomorphic phantoms, representing children at ages 1, 5, and 10 years, at locations corresponding to radiosensitive organs. The phantoms were exposed to mock interventional radiology procedures of the head, chest, and abdomen using posteroanterior and lateral geometries, varying magnification, and fluoroscopy or DSA exposures. Effective doses were calculated from organ doses recorded by the MOSFET dosimeters and are presented in look-up tables according to the different age groups. The largest effective dose burden for fluoroscopy was recorded for posteroanterior and lateral abdominal procedures (0.2-1.1 mSv/min of fluoroscopy), whereas procedures of the head resulted in the lowest effective doses (0.02-0.08 mSv/min of fluoroscopy). DSA exposures of the abdomen imparted higher doses (0.02-0.07 mSv/DSA frame) than did those involving the head and chest. Patient doses during interventional procedures vary significantly depending on the type of procedure. User-friendly look-up tables may provide a helpful tool for health care providers in estimating effective doses for an individual procedure.

  12. Feasibility of three-dimensional magnetic resonance angiography-fluoroscopy image fusion technique in guiding complex endovascular aortic procedures in patients with renal insufficiency.

    PubMed

    Schwein, Adeline; Chinnadurai, Ponraj; Shah, Dipan J; Lumsden, Alan B; Bechara, Carlos F; Bismuth, Jean

    2017-05-01

    Three-dimensional image fusion of preoperative computed tomography (CT) angiography with fluoroscopy using intraoperative noncontrast cone-beam CT (CBCT) has been shown to improve endovascular procedures by reducing procedure length, radiation dose, and contrast media volume. However, patients with a contraindication to CT angiography (renal insufficiency, iodinated contrast allergy) may not benefit from this image fusion technique. The primary objective of this study was to evaluate the feasibility of magnetic resonance angiography (MRA) and fluoroscopy image fusion using noncontrast CBCT as a guidance tool during complex endovascular aortic procedures, especially in patients with renal insufficiency. All endovascular aortic procedures done under MRA image fusion guidance at a single-center were retrospectively reviewed. The patients had moderate to severe renal insufficiency and underwent diagnostic contrast-enhanced magnetic resonance imaging after gadolinium or ferumoxytol injection. Relevant vascular landmarks electronically marked in MRA images were overlaid on real-time two-dimensional fluoroscopy for image guidance, after image fusion with noncontrast intraoperative CBCT. Technical success, time for image registration, procedure time, fluoroscopy time, number of digital subtraction angiography (DSA) acquisitions before stent deployment or vessel catheterization, and renal function before and after the procedure were recorded. The image fusion accuracy was qualitatively evaluated on a binary scale by three physicians after review of image data showing virtual landmarks from MRA on fluoroscopy. Between November 2012 and March 2016, 10 patients underwent endovascular procedures for aortoiliac aneurysmal disease or aortic dissection using MRA image fusion guidance. All procedures were technically successful. A paired t-test analysis showed no difference between preimaging and postoperative renal function (P = .6). The mean time required for MRA-CBCT image fusion was 4:09 ± 01:31 min:sec. Total fluoroscopy time was 20.1 ± 6.9 minutes. Five of 10 patients (50%) underwent stent graft deployment without any predeployment DSA acquisition. Three of six vessels (50%) were cannulated under image fusion guidance without any precannulation DSA runs, and the remaining vessels were cannulated after one planning DSA acquisition. Qualitative evaluation showed 14 of 22 virtual landmarks (63.6%) from MRA overlaid on fluoroscopy were completely accurate, without the need for adjustment. Five of eight incorrect virtual landmarks (iliac and visceral arteries) resulted from vessel deformation caused by endovascular devices. Ferumoxytol or gadolinium-enhanced MRA imaging and image fusion with fluoroscopy using noncontrast CBCT is feasible and allows patients with renal insufficiency to benefit from optimal guidance during complex endovascular aortic procedures, while preserving their residual renal function. Copyright © 2016 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

  13. Learning curve in transradial cardiac catheterization: procedure-related parameters stratified by operators' transradial volume.

    PubMed

    Kasasbeh, Ehab S; Parvez, Babar; Huang, Robert L; Hasselblad, Michele Marie; Glazer, Mark D; Salloum, Joseph G; Cleator, John H; Zhao, David X

    2012-11-01

    To determine whether radial artery access is associated with a reduction in fluoroscopy time, procedure time, and other procedural variables over a 27-month period during which the radial artery approach was incorporated in a single academic Medical Center. Although previous studies have demonstrated a relationship between increased volume and decreased procedural time, no studies have looked at the integration of radial access over time. Data were collected from consecutive patients who presented to the Vanderbilt University Medical Center cardiac catheterization laboratory from January 1, 2009 to April 1, 2011. Patients who underwent radial access diagnostic catheterization with and without percutaneous coronary intervention were included in this study. A total of 1112 diagnostic cardiac catheterizations through the radial access site were analyzed. High-volume, intermediate-volume, and low-volume operators were grouped based on the percentage of procedures performed through a radial approach. From 2009 to 2011, there was a significant decrease in fluoroscopy time in all operator groups for diagnostic catheterization (P=.035). The high-volume operator group had 1.88 and 3.66 minute reductions in fluoroscopy time compared to the intermediate- and low-volume operator groups, respectively (both P<.001). Likewise, the intermediate-volume operator group had a 1.77 minute improvement compared to the low-volume operator group, but this did not reach statistical significance (P=.102). The improvement in fluoroscopy time and other procedure-related parameters was seen after approximately 25 cases with further improvement after 75 cases. The incorporation of the radial access approach in the cardiac catheterization laboratory led to a decrease in fluoroscopy time for each operator and operator group over the last 3 years. Our data demonstrated that higher-volume radial operators have better procedure, room, and fluoroscopy times when compared to intermediate- and low-volume operators. However, lower-volume operators have a reduction in procedure-related parameters with increased radial cases. Number of procedures needed to become sufficient was demonstrated in the current study.

  14. Endoscopic retrograde JJ-stenting of the ureter without fluoroscopy guidance--an appraisal of outcome.

    PubMed

    Shuaibu, S I; Gidado, S; Oseni-Momodu, E

    2013-01-01

    JJ- ureteral stenting is a means of relieving ureteric obstruction. It is done as a retrograde or antegrade procedure, usually under fluoroscopy guidance. We reviewed our results in 2 independent tertiary health centers in Nigeria which lack fluoroscopy units. A 2 year retrospective review of data of patients who had retrograde JJ- ureteric stenting was done. Data relating to age, indication and outcome of procedure were retrieved and analysed. 22 (71%) patients had successful retrograde JJ- ureteric stenting out of 31 patients who were taken for the procedure. These 22 patients had stenting of 27 ureteric units. Mean age was 48.5 years. Commonest indication was carcinoma of the cervix (31.8%). Commonest complication was irritative lower urinary tract symptoms (43.5%). In spite of inherent complications, JJ-stenting is a simple and safe technique. Therefore, the decision to attempt JJ -stenting in carefully selected patients in the absence of fluoroscopy is acceptable.

  15. A simple ergonomic measure reduces fluoroscopy time during ERCP: A multivariate analysis.

    PubMed

    Jowhari, Fahd; Hopman, Wilma M; Hookey, Lawrence

    2017-03-01

    Background and study aims  Endoscopic retrograde cholangiopancreatgraphy (ERCP) carries a radiation risk to patients undergoing the procedure and the team performing it. Fluoroscopy time (FT) has been shown to have a linear relationship with radiation exposure during ERCP. Recent modifications to our ERCP suite design were felt to impact fluoroscopy time and ergonomics. This multivariate analysis was therefore undertaken to investigate these effects, and to identify and validate various clinical, procedural and ergonomic factors influencing the total fluoroscopy time during ERCP. This would better assist clinicians with predicting prolonged fluoroscopic durations and to undertake relevant precautions accordingly. Patients and methods  A retrospective analysis of 299 ERCPs performed by 4 endoscopists over an 18-month period, at a single tertiary care center was conducted. All inpatients/outpatients (121 males, 178 females) undergoing ERCP for any clinical indication from January 2012 to June 2013 in the chosen ERCP suite were included in the study. Various predetermined clinical, procedural and ergonomic factors were obtained via chart review. Univariate analyses identified factors to be included in the multivariate regression model with FT as the dependent variable. Results  Bringing the endoscopy and fluoroscopy screens next to each other was associated with a significantly lesser FT than when the screens were separated further (-1.4 min, P  = 0.026). Other significant factors associated with a prolonged FT included having a prior ERCP (+ 1.4 min, P  = 0.031), and more difficult procedures (+ 4.2 min for each level of difficulty, P  < 0.001). ERCPs performed by high-volume endoscopists used lesser FT vs. low-volume endoscopists (-1.82, P = 0.015). Conclusions  Our study has identified and validated various factors that affect the total fluoroscopy time during ERCP. This is the first study to show that decreasing the distance between the endoscopy and fluoroscopy screens in the ERCP suite significantly reduces the total fluoroscopy time, and therefore radiation exposure to patients and staff involved in the procedure.

  16. Laser Guidance in C-Arm Cone-Beam CT-Guided Radiofrequency Ablation of Osteoid Osteoma Reduces Fluoroscopy Time

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kroes, Maarten W., E-mail: Maarten.Kroes@radboudumc.nl; Busser, Wendy M. H.; Hoogeveen, Yvonne L.

    PurposeTo assess whether laser guidance can reduce fluoroscopy and procedure time of cone-beam computed tomography (CBCT)-guided radiofrequency (RF) ablations of osteoid osteoma compared to freehand CBCT guidance.Materials and Methods32 RF ablations were retrospectively analyzed, 17 laser-guided and 15 procedures using the freehand technique. Subgroup selection of 18 ablations in the hip–pelvic region with a similar degree of difficulty was used for a direct comparison. Data are presented as median (ranges).ResultsComparison of all 32 ablations resulted in fluoroscopy times of 365 s (193–878 s) for freehand and 186 s (75–587 s) for laser-guided procedures (p = 0.004). Corresponding procedure times were 56 min (35–97 min) and 52 min (30–85 min) (p = 0.355).more » The subgroup showed comparable target sizes, needle path lengths, and number of scans between groups. Fluoroscopy times were lower for laser-guided procedures, 215 s (75–413 s), compared to 384 s (193–878 s) for freehand (p = 0.012). Procedure times were comparable between groups, 51 min (30–72 min) for laser guidance and 58 min (35–79 min) for freehand (p = 0.172).ConclusionAdding laser guidance to CBCT-guided osteoid osteoma RF ablations significantly reduced fluoroscopy time without increasing procedure time.Level of EvidenceLevel 4, case series.« less

  17. Comparison of Ultrasound-Guided and Fluoroscopy-Assisted Antegrade Common Femoral Artery Puncture Techniques

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Slattery, Michael M.; Goh, Gerard S.; Power, Sarah

    PurposeTo prospectively compare the procedural time and complication rates of ultrasound-guided and fluoroscopy-assisted antegrade common femoral artery (CFA) puncture techniques.Materials and MethodsHundred consecutive patients, undergoing a vascular procedure for which an antegrade approach was deemed necessary/desirable, were randomly assigned to undergo either ultrasound-guided or fluoroscopy-assisted CFA puncture. Time taken from administration of local anaesthetic to vascular sheath insertion in the superficial femoral artery (SFA), patients’ age, body mass index (BMI), fluoroscopy radiation dose, haemostasis method and immediate complications were recorded. Mean and median values were calculated and statistically analysed with unpaired t tests.ResultsSixty-nine male and 31 female patients underwent antegrademore » puncture (mean age 66.7 years). The mean BMI was 25.7 for the ultrasound-guided (n = 53) and 25.3 for the fluoroscopy-assisted (n = 47) groups. The mean time taken for the ultrasound-guided puncture was 7 min 46 s and for the fluoroscopy-assisted technique was 9 min 41 s (p = 0.021). Mean fluoroscopy dose area product in the fluoroscopy group was 199 cGy cm{sup 2}. Complications included two groin haematomas in the ultrasound-guided group and two retroperitoneal haematomas and one direct SFA puncture in the fluoroscopy-assisted group.ConclusionUltrasound-guided technique is faster and safer for antegrade CFA puncture when compared to the fluoroscopic-assisted technique alone.« less

  18. Technical results and effects of operator experience on uterine artery embolization for fibroids: the Ontario Uterine Fibroid Embolization Trial.

    PubMed

    Pron, Gaylene; Bennett, John; Common, Andrew; Sniderman, Kenneth; Asch, Murray; Bell, Stuart; Kozak, Roman; Vanderburgh, Leslie; Garvin, Greg; Simons, Martin; Tran, Cuong; Kachura, John

    2003-05-01

    To document the technical results and spectrum of practice of uterine artery embolization (UAE) for fibroids in the health care setting in Canada. The effects of interventional radiologist's (IR's) experience with UAE on procedure and fluoroscopy time were also investigated. The study involved a multicenter prospective single-arm clinical treatment trial and included the practices of 11 IRs at eight university-affiliated teaching and community hospitals. Vascular access with percutaneous femoral artery approach was followed by transcatheter delivery of polyvinyl alcohol (PVA) particles into uterine arteries with fluoroscopic guidance. Technical success, complications, procedural time, fluoroscopy time, and effects of operator experience were outcomes analyzed. Between November 1998 and November 2000, 570 embolization procedures were performed in 555 patients. UAE was bilaterally successful in 97% (95% CI: 95%-98%). Variant anatomy was the most common reason for failure to embolize bilaterally. The procedural complication rate was 5.3% (95% CI: 3.6%-7.4%). Of the 30 events, three involved major complications (one seizure and two allergic reactions) that resulted in additional care or extended hospital stay. Procedure time and fluoroscopy time averaged 61 minutes (95% CI; 58-63 minutes) and 18.9 minutes (95% CI; 18-19.8) and varied significantly among IRs (P <.001; P <.001). The average 27% reduction in procedure time (20 minutes; P <.001) and 24% reduction in fluoroscopy time (5.1 minutes; P <.001) with increasing UAE experience were significant. A high level of technical success with few complications was obtained with a variety of operators in diverse practice settings. Increased experience in UAE significantly reduced procedure and fluoroscopy time.

  19. SU-F-I-77: Radiation Dose in Cardiac Catheterization Procedures: Impact of a Systematic Reduction in Pulsed Fluoroscopy Frame Rate

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Schultz, C; Dixon, S

    Purpose: To evaluate whether one small systematic reduction in fluoroscopy frame rate has a significant effect on the total air kerma and/or dose area product for diagnostic and interventional cardiac catheterization procedures. Methods: The default fluoroscopy frame rate (FFR) was lowered from 15 to 10 fps in 5 Siemens™ Axiom Artis cardiac catheterization labs (CCL) on July 1, 2013. A total of 7212 consecutive diagnostic and interventional CCL procedures were divided into two study groups: 3602 procedures from 10/1/12 –6/30/13 with FFR of 15 fps; and 3610 procedures 7/1/13 – 3/31/14 at 10 fps. For each procedure, total air kermamore » (TAK), fluoroscopy skin dose (FSD), total/fluoroscopy dose area products (TAD, FAD), and total fluoroscopy time (FT) were recorded. Patient specific data collected for each procedure included: BSA, sex, height, weight, interventional versus diagnostic; and elective versus emergent. Results: For pre to post change in FFR, each categorical variable was compared using Pearson’s Chi-square test, Odds ratios and 95% confidence intervals. No statistically significant difference in BSA, height, weight, number of interventional versus diagnostic, elective versus emergent procedures was found between the two study groups. Decreasing the default FFR from 15 fps to 10 fps in the two study groups significantly reduced TAK from 1305 to 1061 mGy (p<0.0001), FSD from 627 to 454 mGy (p<0.0001), TAD from 8681 to 6991 uGy × m{sup 2}(p<0.0001), and FAD from 4493 to 3297 uGy × m{sup 2}(p<0.0001). No statistically significant difference in FT was noted. Clinical image quality was not analyzed, and reports of noticeable effects were minimal. From July 1, 2013 to date, the default FFR has remained 10 fps. Conclusion: Reducing the FFR from 15 to 10 fps significantly reduced total air kerma and dose area product which may decrease risk for potential radiation-induced skin injuries and improve patient outcomes.« less

  20. Oral versus intravenous premedication for small bowel biopsy in children: effect on procedure and fluoroscopy times.

    PubMed

    Stenhammar, L; Wärngård, O; Lewander, P; Nordvall, M

    1993-01-01

    Oral alimemazine and cisapride, or diazepam and cisapride, or iv midazolam and metoclopramide were given as premedication for small bowel biopsy to three groups of children from a total population of 185 individuals. The biopsy procedures were performed under intermittent fluoroscopy and times for both were recorded. The median biopsy procedure time was significantly shorter in children given iv midazolam and metoclopramide (6 min) compared to those given oral premedication (10 min) (p < 0.001). The median fluoroscopy time was very short in all groups, ranging between 3 and 6 s. It is concluded that iv premedication is superior to oral premedication for small bowel biopsy in children because more effective sedation is obtained.

  1. Navigation for fluoroscopy-guided cryo-balloon ablation procedures of atrial fibrillation

    NASA Astrophysics Data System (ADS)

    Bourier, Felix; Brost, Alexander; Kleinoeder, Andreas; Kurzendorfer, Tanja; Koch, Martin; Kiraly, Attila; Schneider, Hans-Juergen; Hornegger, Joachim; Strobel, Norbert; Kurzidim, Klaus

    2012-02-01

    Atrial fibrillation (AFib), the most common arrhythmia, has been identified as a major cause of stroke. The current standard in interventional treatment of AFib is the pulmonary vein isolation (PVI). PVI is guided by fluoroscopy or non-fluoroscopic electro-anatomic mapping systems (EAMS). Either classic point-to-point radio-frequency (RF)- catheter ablation or so-called single-shot-devices like cryo-balloons are used to achieve electrically isolation of the pulmonary veins and the left atrium (LA). Fluoroscopy-based systems render overlay images from pre-operative 3-D data sets which are then merged with fluoroscopic imaging, thereby adding detailed 3-D information to conventional fluoroscopy. EAMS provide tracking and visualization of RF catheters by means of electro-magnetic tracking. Unfortunately, current navigation systems, fluoroscopy-based or EAMS, do not provide tools to localize and visualize single shot devices like cryo-balloon catheters in 3-D. We present a prototype software for fluoroscopy-guided ablation procedures that is capable of superimposing 3-D datasets as well as reconstructing cyro-balloon catheters in 3-D. The 3-D cyro-balloon reconstruction was evaluated on 9 clinical data sets, yielded a reprojected 2-D error of 1.72 mm +/- 1.02 mm.

  2. CTA with fluoroscopy image fusion guidance in endovascular complex aortic aneurysm repair.

    PubMed

    Sailer, A M; de Haan, M W; Peppelenbosch, A G; Jacobs, M J; Wildberger, J E; Schurink, G W H

    2014-04-01

    To evaluate the effect of intraoperative guidance by means of live fluoroscopy image fusion with computed tomography angiography (CTA) on iodinated contrast material volume, procedure time, and fluoroscopy time in endovascular thoraco-abdominal aortic repair. CTA with fluoroscopy image fusion road-mapping was prospectively evaluated in patients with complex aortic aneurysms who underwent fenestrated and/or branched endovascular repair (FEVAR/BEVAR). Total iodinated contrast material volume, overall procedure time, and fluoroscopy time were compared between the fusion group (n = 31) and case controls (n = 31). Reasons for potential fusion image inaccuracy were analyzed. Fusion imaging was feasible in all patients. Fusion image road-mapping was used for navigation and positioning of the devices and catheter guidance during access to target vessels. Iodinated contrast material volume and procedure time were significantly lower in the fusion group than in case controls (159 mL [95% CI 132-186 mL] vs. 199 mL [95% CI 170-229 mL], p = .037 and 5.2 hours [95% CI 4.5-5.9 hours] vs. 6.3 hours (95% CI 5.4-7.2 hours), p = .022). No significant differences in fluoroscopy time were observed (p = .38). Respiration-related vessel displacement, vessel elongation, and displacement by stiff devices as well as patient movement were identified as reasons for fusion image inaccuracy. Image fusion guidance provides added value in complex endovascular interventions. The technology significantly reduces iodinated contrast material dose and procedure time. Copyright © 2014 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

  3. Effective doses in children: association with common complex imaging techniques used during interventional radiology procedures.

    PubMed

    Lai, Priscilla; McNeil, Sarah M; Gordon, Christopher L; Connolly, Bairbre L

    2014-12-01

    The purpose of this study was to determine the range of effective doses associated with imaging techniques used during interventional radiology procedures on children. A pediatric phantom set (1, 5, and 10 years) coupled with high-sensitivity metal oxide semiconductor field effect transistor (MOSFET) dosimeters was used to calculate effective doses. Twenty MOSFETs were inserted into each phantom at radiosensitive organ locations. The phantoms were exposed to mock head, chest, and abdominal interventional radiology procedures performed with different geometries and magnifications. Fluoroscopy, digital subtraction angiography (DSA), and spin angiography were simulated on each phantom. Road mapping was conducted only on the 5-year-old phantom. International Commission on Radiological Protection publication 103 tissue weights were applied to the organ doses recorded with the MOSFETs to determine effective dose. For easy application to clinical cases, doses were normalized per minute of fluoroscopy and per 10 frames of DSA or spin angiography. Effective doses from DSA, angiography, and fluoroscopy were higher for younger ages because of magnification use and were largest for abdominal procedures. DSA of the head, chest, and abdomen (normalized per 10 frames) imparted doses 2-3 times as high as corresponding doses per minute of fluoroscopy while all other factors remained unchanged (age, projection, collimation, magnification). Three to five frames of DSA imparted an effective dose equal to doses from 1 minute of fluoroscopy. Doses from spin angiography were almost one-half the doses received from an equivalent number of frames of DSA. Patient effective doses during interventional procedures vary substantially depending on procedure type but tend to be higher because of magnification use in younger children and higher in the abdomen.

  4. The Use of Laser Guidance Reduces Fluoroscopy Time for C-Arm Cone-Beam Computed Tomography-Guided Biopsies

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kroes, Maarten W., E-mail: Maarten.Kroes@radboudumc.nl; Strijen, Marco J. L. van, E-mail: m.van.strijen@antoniusziekenhuis.nl; Braak, Sicco J., E-mail: sjbraak@gmail.com

    2016-09-15

    PurposeWhen using laser guidance for cone-beam computed tomography (CBCT)-guided needle interventions, planned needle paths are visualized to the operator without the need to switch between entry- and progress-view during needle placement. The current study assesses the effect of laser guidance during CBCT-guided biopsies on fluoroscopy and procedure times.Materials and MethodsProspective data from 15 CBCT-guided biopsies of 8–65 mm thoracic and abdominal lesions assisted by a ceiling-mounted laser guidance technique were compared to retrospective data of 36 performed CBCT-guided biopsies of lesions >20 mm using the freehand technique. Fluoroscopy time, procedure time, and number of CBCT-scans were recorded. All data are presented asmore » median (ranges).ResultsFor biopsies using the freehand technique, more fluoroscopy time was necessary to guide the needle onto the target, 165 s (83–333 s) compared to 87 s (44–190 s) for laser guidance (p < 0.001). Procedure times were shorter for freehand-guided biopsies, 24 min versus 30 min for laser guidance (p < 0.001).ConclusionThe use of laser guidance during CBCT-guided biopsies significantly reduces fluoroscopy time.« less

  5. Magnetic navigation system for percutaneous coronary intervention

    PubMed Central

    Qi, Zhiyong; Wu, Bangwei; Luo, Xinping; Zhu, Jun; Shi, Haiming; Jin, Bo

    2016-01-01

    Abstract Background: Magnetic navigation system (MNS) allows calculation of the vessel coordinates in real space within the patient's chest for percutaneous coronary intervention (PCI). However, its impact on the procedural parameters and clinical outcomes is still a matter of debate. To derive a more precise estimation of the relationship, a meta-analysis was performed. Methods and Results: Studies exploring the advantages of MNS were identified in English-language articles by search of Medline, Web of Science, and Cochrane Library Databases (inception to October 2015). A standardized protocol was used to extract details on study design, region origin, demographic data, lesion type, and clinical outcomes. The main outcome measures were contrast consumption, procedural success rate, contrast used for wire crossing, procedure time to cross the lesions, and the fluoroscopy time fluoroscopy time. A total of 12 clinical trials involving 2174 patients were included for analysis (902 patients in the magnetic PCI group and 1272 in the conventional PCI group). Overall, contrast consumption was decreased by 40.45 mL (95% confidence interval [CI] −70.98 to −9.92, P = 0.009) in magnetic PCI group compared with control group. In addition, magnetic PCI was associated with significantly decreasing procedural time by 2.17 minutes (95% CI −3.91 to −0.44, P = 0.01) and the total fluoroscopy time was significantly decreased by 1.43 minutes (95% CI −2.29 to −0.57, P = 0.001) in magnetic PCI group. However, procedural success rate, contrast used for wire crossing, procedure time to cross the lesions, and the fluoroscopy time to cross the lesions demonstrated that no statistically difference was observed between 2 groups. Conclusion: The present meta-analysis indicated an improvement of overall contrast consumption, total procedural time, and fluoroscopy time in magnetic PCI group. However, no significant advantages were observed associated with procedural success rate. PMID:27442645

  6. Magnetic navigation system for percutaneous coronary intervention: A meta-analysis.

    PubMed

    Qi, Zhiyong; Wu, Bangwei; Luo, Xinping; Zhu, Jun; Shi, Haiming; Jin, Bo

    2016-07-01

    Magnetic navigation system (MNS) allows calculation of the vessel coordinates in real space within the patient's chest for percutaneous coronary intervention (PCI). However, its impact on the procedural parameters and clinical outcomes is still a matter of debate. To derive a more precise estimation of the relationship, a meta-analysis was performed. Studies exploring the advantages of MNS were identified in English-language articles by search of Medline, Web of Science, and Cochrane Library Databases (inception to October 2015). A standardized protocol was used to extract details on study design, region origin, demographic data, lesion type, and clinical outcomes. The main outcome measures were contrast consumption, procedural success rate, contrast used for wire crossing, procedure time to cross the lesions, and the fluoroscopy time fluoroscopy time. A total of 12 clinical trials involving 2174 patients were included for analysis (902 patients in the magnetic PCI group and 1272 in the conventional PCI group). Overall, contrast consumption was decreased by 40.45 mL (95% confidence interval [CI] -70.98 to -9.92, P = 0.009) in magnetic PCI group compared with control group. In addition, magnetic PCI was associated with significantly decreasing procedural time by 2.17 minutes (95% CI -3.91 to -0.44, P = 0.01) and the total fluoroscopy time was significantly decreased by 1.43 minutes (95% CI -2.29 to -0.57, P = 0.001) in magnetic PCI group. However, procedural success rate, contrast used for wire crossing, procedure time to cross the lesions, and the fluoroscopy time to cross the lesions demonstrated that no statistically difference was observed between 2 groups. The present meta-analysis indicated an improvement of overall contrast consumption, total procedural time, and fluoroscopy time in magnetic PCI group. However, no significant advantages were observed associated with procedural success rate.

  7. CT fluoroscopy-guided preoperative short hook wire placement for small pulmonary lesions: evaluation of safety and identification of risk factors for pneumothorax.

    PubMed

    Iguchi, Toshihiro; Hiraki, Takao; Gobara, Hideo; Fujiwara, Hiroyasu; Matsui, Yusuke; Miyoshi, Shinichiro; Kanazawa, Susumu

    2016-01-01

    To retrospectively evaluate the safety of computed tomography (CT) fluoroscopy-guided short hook wire placement for video-assisted thoracoscopic surgery and the risk factors for pneumothorax associated with this procedure. We analyzed 267 short hook wire placements for 267 pulmonary lesions (mean diameter, 9.9 mm). Multiple variables related to the patients, lesions, and procedures were assessed to determine the risk factors for pneumothorax. Complications (219 grade 1 and 4 grade 2 adverse events) occurred in 196 procedures. No grade 3 or above adverse events were observed. Univariate analysis revealed increased vital capacity (odds ratio [OR], 1.518; P = 0.021), lower lobe lesion (OR, 2.343; P =0.001), solid lesion (OR, 1.845; P = 0.014), prone positioning (OR, 1.793; P = 0.021), transfissural approach (OR, 11.941; P = 0.017), and longer procedure time (OR, 1.036; P = 0.038) were significant predictors of pneumothorax. Multivariate analysis revealed only the transfissural approach (OR, 12.171; P = 0.018) and a longer procedure time (OR, 1.048; P = 0.012) as significant independent predictors. Complications related to CT fluoroscopy-guided preoperative short hook wire placement often occurred, but all complications were minor. A transfissural approach and longer procedure time were significant independent predictors of pneumothorax. Complications related to CT fluoroscopy-guided preoperative short hook wire placement often occur. Complications are usually minor and asymptomatic. A transfissural approach and longer procedure time are significant independent predictors of pneumothorax.

  8. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Das, Raj, E-mail: rajdas@nhs.net, E-mail: raj.das@stgeorges.nhs.uk; Lucatelli, Pierleone, E-mail: pierleone.lucatelli@gmail.com; Wang, Haofan, E-mail: wwhhff123@gmail.com

    AimA clear understanding of operator experience is important in improving technical success whilst minimising patient risk undergoing endovascular procedures, and there is the need to ensure that trainees have the appropriate skills as primary operators. The aim of the study is to retrospectively analyse uterine artery embolisation (UAE) procedures performed by interventional radiology (IR) trainees at an IR training unit analysing fluoroscopy times and radiation dose as surrogate markers of technical skill.MethodsTen IR fellows were primary operator in 200 UAE procedures over a 5-year period. We compared fluoroscopy times, radiation dose and complications, after having them categorised according to threemore » groups: Group 1, initial five, Group 2, >5 procedures and Group 3, penultimate five UAE procedures. We documented factors that may affect screening time (number of vials employed and use of microcatheters).ResultsMean fluoroscopy time was 18.4 (±8.1), 17.3 (±9.0), 16.3 (±8.4) min in Groups 1, 2 and 3, respectively. There was no statistically significant difference between these groups (p > 0.05) with respect to fluoroscopy time or radiation dose. Analysis after correction for confounding factors showed no statistical significance (p > 0.05). All procedures were technically successful, and total complication rate was 4 %.ConclusionUAE was chosen as a highly standardised procedure followed by IR practitioners. Although there is a non-significant trend for shorter screening times with experience, technical success and safety were not compromised with appropriate Consultant supervision, which illustrates a safe construct for IR training. This is important and reassuring information for patients undergoing a procedure in a training unit.« less

  9. Non-fluoroscopic navigation systems for radiofrequency catheter ablation for supraventricular tachycardia reduce ionising radiation exposure.

    PubMed

    See, Jason; Amora, Jonah L; Lee, Sheldon; Lim, Paul; Teo, Wee Siong; Tan, Boon Yew; Ho, Kah Leng; Lee, Chee Wan; Ching, Chi-Keong

    2016-07-01

    The use of non-fluoroscopic systems (NFS) to guide radiofrequency catheter ablation (RFCA) for the treatment of supraventricular tachycardia (SVT) is associated with lower radiation exposure. This study aimed to determine if NFS reduces fluoroscopy time, radiation dose and procedure time. We prospectively enrolled patients undergoing RFCA for SVT. NFS included EnSiteTM NavXTM or CARTO® mapping. We compared procedure and fluoroscopy times, and radiation exposure between NFS and conventional fluoroscopy (CF) cohorts. Procedural success, complications and one-year success rates were reported. A total of 200 patients over 27 months were included and RFCA was guided by NFS for 79 patients; those with atrioventricular nodal reentrant tachycardia (AVNRT), left-sided atrioventricular reentrant tachycardia (AVRT) and right-sided AVRT were included (n = 101, 63 and 36, respectively). Fluoroscopy times were significantly lower with NFS than with CF (10.8 ± 11.1 minutes vs. 32.0 ± 27.5 minutes; p < 0.001). The mean fluoroscopic dose area product was also significantly reduced with NFS (NSF: 5,382 ± 5,768 mGy*cm2 vs. CF: 21,070 ± 23,311 mGy*cm2; p < 0.001); for all SVT subtypes. There was no significant reduction in procedure time, except for left-sided AVRT ablation (NFS: 79.2 minutes vs. CF: 116.4 minutes; p = 0.001). Procedural success rates were comparable (NFS: 97.5% vs. CF: 98.3%) and at one-year follow-up, there was no significant difference in the recurrence rates (NFS: 5.2% vs. CF: 4.2%). No clinically significant complications were observed in both groups. The use of NFS for RFCA for SVT is safe, with significantly reduced radiation dose and fluoroscopy time. Copyright © Singapore Medical Association.

  10. Sensitivity of the diagnostic radiological index of protection to procedural factors in fluoroscopy.

    PubMed

    Jones, A Kyle; Pasciak, Alexander S; Wagner, Louis K

    2016-07-01

    To evaluate the sensitivity of the diagnostic radiological index of protection (DRIP), used to quantify the protective value of radioprotective garments, to procedural factors in fluoroscopy in an effort to determine an appropriate set of scatter-mimicking primary beams to be used in measuring the DRIP. Monte Carlo simulations were performed to determine the shape of the scattered x-ray spectra incident on the operator in different clinical fluoroscopy scenarios, including interventional radiology and interventional cardiology (IC). Two clinical simulations studied the sensitivity of the scattered spectrum to gantry angle and patient size, while technical factors were varied according to measured automatic dose rate control (ADRC) data. Factorial simulations studied the sensitivity of the scattered spectrum to gantry angle, field of view, patient size, and beam quality for constant technical factors. Average energy (Eavg) was the figure of merit used to condense fluence in each energy bin to a single numerical index. Beam quality had the strongest influence on the scattered spectrum in fluoroscopy. Many procedural factors affect the scattered spectrum indirectly through their effect on primary beam quality through ADRC, e.g., gantry angle and patient size. Lateral C-arm rotation, common in IC, increased the energy of the scattered spectrum, regardless of the direction of rotation. The effect of patient size on scattered radiation depended on ADRC characteristics, patient size, and procedure type. The scattered spectrum striking the operator in fluoroscopy is most strongly influenced by primary beam quality, particularly kV. Use cases for protective garments should be classified by typical procedural primary beam qualities, which are governed by the ADRC according to the impacts of patient size, anatomical location, and gantry angle.

  11. First metatarsal closing base wedge osteotomy using real-time fluoroscopy.

    PubMed

    Toepp, F C; Salcedo, M

    1991-01-01

    A minimal incision surgery approach to metatarsus primus adductus is presented. The percutaneous closing base wedge osteotomy is performed using real-time intraoperative fluoroscopy. The advantages and disadvantages of this minimal incision surgical procedure are discussed.

  12. Emergency endotracheal intubation under fluoroscopy guidance for patients with acute dyspnea or asphyxia.

    PubMed

    Jiao, Dechao; Xie, Na; Han, Xinwei; Wu, Gang

    2016-11-01

    To evaluate the feasibility and effectiveness of emergency endotracheal intubation (EEI) under fluoroscopy guidance for patients with acute dyspnea or asphyxia. From October 2011 to October 2014, of 1521 patients with acute dyspnea or asphyxia who required EEI in 6 departments, 43 patients who experienced intubation difficulty or failure were entered into this study. Data on technical success, procedure time, complications, and clinical outcome were collected. The pulse oxygen saturation and Hugh-Jones classification changes were analyzed. Fluoroscopy-guided EEI was technically successful in all patients. Acute dyspnea had resolved in all patients with clinical success rate 100% after the procedure. There were no serious complications during or after the procedure. The pulse oxygen saturation and Hugh-Jones classification showed significant increase after EEI (P < .05). Further treatments, including tracheal stents (n = 21), surgical resection (n = 16), palliative tracheotomy (n = 4), and bronchoscopic treatment (n = 2), were performed 1 to 72 hours after EEI. During a mean follow-up period of 13.2 months, 13 patients had died and 30 patients remained alive without dyspnea. Fluoroscopy-guided EEI is a safe and feasible procedure, and may serve as an alternative treatment option for patients when traditional EEI is unsuccessful. Copyright © 2016 Elsevier Inc. All rights reserved.

  13. Radiation exposure and contrast agent use related to radial versus femoral arterial access during percutaneous coronary intervention (PCI)-Results of the FERARI study.

    PubMed

    Becher, Tobias; Behnes, Michael; Ünsal, Melike; Baumann, Stefan; El-Battrawy, Ibrahim; Fastner, Christian; Kuschyk, Jürgen; Papavassiliu, Theano; Hoffmann, Ursula; Mashayekhi, Kambis; Borggrefe, Martin; Akin, Ibrahim

    2016-12-01

    Data regarding radiation exposure related to radial versus femoral arterial access in patients undergoing percutaneous coronary intervention (PCI) remain controversial. This study aims to evaluate patients enrolled in the FERARI study regarding radiation exposure, fluoroscopy time and contrast agent use. The Femoral Closure versus Radial Compression Devices Related to Percutaneous Coronary Interventions (FERARI) study evaluated prospectively 400 patients between February 2014 and May 2015 undergoing PCI either using the radial or femoral access. In these 400 patients, baseline characteristics, procedural data such as procedural duration, fluoroscopy time, dose-area product (DAP) as well as the amount of contrast agent used were documented and analyzed. Median fluoroscopy time was not significantly different in patients undergoing radial versus femoral access (12.2 vs. 9.8min, p=0.507). Furthermore, median DAP (54.5 vs. 52.0 Gycm2, p=0.826), procedural duration (46.0 vs. 45.0min, p=0.363) and contrast agent use (185.5 vs. 199.5ml, p=0.742) were also similar in radial and femoral PCI. There was no difference regarding median fluoroscopy time, procedural duration, radiation dose or contrast agent use between radial versus femoral arterial access in PCI. Copyright © 2016 Elsevier Inc. All rights reserved.

  14. CT fluoroscopy-guided renal tumour cutting needle biopsy: retrospective evaluation of diagnostic yield, safety, and risk factors for diagnostic failure.

    PubMed

    Iguchi, Toshihiro; Hiraki, Takao; Matsui, Yusuke; Fujiwara, Hiroyasu; Sakurai, Jun; Masaoka, Yoshihisa; Gobara, Hideo; Kanazawa, Susumu

    2018-01-01

    To evaluate retrospectively the diagnostic yield, safety, and risk factors for diagnostic failure of computed tomography (CT) fluoroscopy-guided renal tumour biopsy. Biopsies were performed for 208 tumours (mean diameter 2.3 cm; median diameter 2.1 cm; range 0.9-8.5 cm) in 199 patients. One hundred and ninety-nine tumours were ≤4 cm. All 208 initial procedures were divided into diagnostic success and failure groups. Multiple variables related to the patients, lesions, and procedures were assessed to determine the risk factors for diagnostic failure. After performing 208 initial and nine repeat biopsies, 180 malignancies and 15 benign tumours were pathologically diagnosed, whereas 13 were not diagnosed. In 117 procedures, 118 Grade I and one Grade IIIa adverse events (AEs) occurred. Neither Grade ≥IIIb AEs nor tumour seeding were observed within a median follow-up period of 13.7 months. Logistic regression analysis revealed only small tumour size (≤1.5 cm; odds ratio 3.750; 95% confidence interval 1.362-10.326; P = 0.011) to be a significant risk factor for diagnostic failure. CT fluoroscopy-guided renal tumour biopsy is a safe procedure with a high diagnostic yield. A small tumour size (≤1.5 cm) is a significant risk factor for diagnostic failure. • CT fluoroscopy-guided renal tumour biopsy has a high diagnostic yield. • CT fluoroscopy-guided renal tumour biopsy is safe. • Small tumour size (≤1.5 cm) is a risk factor for diagnostic failure.

  15. CT fluoroscopy-assisted puncture of thoracic and abdominal masses: a randomized trial.

    PubMed

    Kirchner, Johannes; Kickuth, Ralph; Laufer, Ulf; Schilling, Esther Maria; Adams, Stephan; Liermann, Dieter

    2002-03-01

    We investigated the benefit of real-time guidance of interventional punctures by means of computed tomography fluoroscopy (CTF) compared with the conventional sequential acquisition guidance. In a prospective randomized trial, 75 patients underwent either CTF-guided (group A, n = 50) or sequential CT-guided (group B, n = 25) punctures of thoracic (n = 29) or abdominal (n = 46) masses. CTF was performed on the CT machine (Somatom Plus 4 Power, Siemens Corp., Forchheim, Germany) equipped with the C.A.R.E. Vision application (tube voltage 120 kV, tube current 50 mA, rotational time 0.75 s, slice thickness 10 mm, 8 frames/s). The average procedure time showed a statistically significant difference between the two study groups (group A: 564 s, group B 795 s, P = 0.0032). The mean total mAs was 7089 mAs for the CTF and 4856 mAs for the sequential image-guided intervention, respectively. The sensitivity was 71% specificity 100% positive predictive value 100% and negative predictive value 60% for the CTF-guided puncture, and 68, 100, 100 and 50% for sequential CT, respectively. CTF guidance realizes a time-saving but increases the radiation exposure dosage.

  16. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ierardi, Anna Maria; Duka, Ejona; Radaelli, Alessandro

    AimTo evaluate the feasibility of image fusion (IF) of pre-procedural arterial-phase CT angiography or MR angiography with intra-procedural fluoroscopy for road-mapping in endovascular treatment of aorto-iliac steno-occlusive disease.Materials and MethodsBetween September and November, 2014, we prospectively evaluated 5 patients with chronic aorto-iliac steno-occlusive disease, who underwent endovascular treatment in the angiography suite. Fusion image road-mapping was performed using angiographic phase CT images or MR images acquired before and intra-procedural unenhanced cone-beam CT. Radiation dose of the procedure, volume of intra-procedural iodinated contrast medium, fluoroscopy time, and overall procedural time were recorded. Reasons for potential fusion imaging inaccuracies were also evaluated.ResultsImagemore » co-registration and fusion guidance were feasible in all procedures. Mean radiation dose of the procedure was 60.21 Gycm2 (range 55.02–63.75 Gycm2). The mean total procedure time was 32.2 min (range 27–38 min). The mean fluoroscopy time was 12 min and 3 s. The mean procedural iodinated contrast material dose was 24 mL (range 20–40 mL).ConclusionsIF gives Interventional Radiologists the opportunity to use new technologies in order to improve outcomes with a significant reduction of contrast media administration.« less

  17. Reduction of radiation exposure while maintaining high-quality fluoroscopic images during interventional cardiology using novel x-ray tube technology with extra beam filtering.

    PubMed

    den Boer, A; de Feyter, P J; Hummel, W A; Keane, D; Roelandt, J R

    1994-06-01

    Radiographic technology plays an integral role in interventional cardiology. The number of interventions continues to increase, and the associated radiation exposure to patients and personnel is of major concern. This study was undertaken to determine whether a newly developed x-ray tube deploying grid-switched pulsed fluoroscopy and extra beam filtering can achieve a reduction in radiation exposure while maintaining fluoroscopic images of high quality. Three fluoroscopic techniques were compared: continuous fluoroscopy, pulsed fluoroscopy, and a newly developed high-output pulsed fluoroscopy with extra filtering. To ascertain differences in the quality of images and to determine differences in patient entrance and investigator radiation exposure, the radiated volume curve was measured to determine the required high voltage levels (kVpeak) for different object sizes for each fluoroscopic mode. The fluoroscopic data of 124 patient procedures were combined. The data were analyzed for radiographic projections, image intensifier field size, and x-ray tube kilovoltage levels (kVpeak). On the basis of this analysis, a reference procedure was constructed. The reference procedure was tested on a phantom or dummy patient by all three fluoroscopic modes. The phantom was so designed that the kilovoltage requirements for each projection were comparable to those needed for the average patient. Radiation exposure of the operator and patient was measured during each mode. The patient entrance dose was measured in air, and the operator dose was measured by 18 dosimeters on a dummy operator. Pulsed compared with continuous fluoroscopy could be performed with improved image quality at lower kilovoltages. The patient entrance dose was reduced by 21% and the operator dose by 54%. High-output pulsed fluoroscopy with extra beam filtering compared with continuous fluoroscopy improved the image quality, lowered the kilovoltage requirements, and reduced the patient entrance dose by 55% and the operator dose by 69%. High-output pulsed fluoroscopy with a grid-switched tube and extra filtering improves the image quality and significantly reduces both the operator dose and patient dose.

  18. C-arm positioning using virtual fluoroscopy for image-guided surgery

    NASA Astrophysics Data System (ADS)

    de Silva, T.; Punnoose, J.; Uneri, A.; Goerres, J.; Jacobson, M.; Ketcha, M. D.; Manbachi, A.; Vogt, S.; Kleinszig, G.; Khanna, A. J.; Wolinsky, J.-P.; Osgood, G.; Siewerdsen, J. H.

    2017-03-01

    Introduction: Fluoroscopically guided procedures often involve repeated acquisitions for C-arm positioning at the cost of radiation exposure and time in the operating room. A virtual fluoroscopy system is reported with the potential of reducing dose and time spent in C-arm positioning, utilizing three key advances: robust 3D-2D registration to a preoperative CT; real-time forward projection on GPU; and a motorized mobile C-arm with encoder feedback on C-arm orientation. Method: Geometric calibration of the C-arm was performed offline in two rotational directions (orbit α, orbit β). Patient registration was performed using image-based 3D-2D registration with an initially acquired radiograph of the patient. This approach for patient registration eliminated the requirement for external tracking devices inside the operating room, allowing virtual fluoroscopy using commonly available systems in fluoroscopically guided procedures within standard surgical workflow. Geometric accuracy was evaluated in terms of projection distance error (PDE) in anatomical fiducials. A pilot study was conducted to evaluate the utility of virtual fluoroscopy to aid C-arm positioning in image guided surgery, assessing potential improvements in time, dose, and agreement between the virtual and desired view. Results: The overall geometric accuracy of DRRs in comparison to the actual radiographs at various C-arm positions was PDE (mean ± std) = 1.6 ± 1.1 mm. The conventional approach required on average 8.0 ± 4.5 radiographs spent "fluoro hunting" to obtain the desired view. Positioning accuracy improved from 2.6o ± 2.3o (in α) and 4.1o ± 5.1o (in β) in the conventional approach to 1.5o ± 1.3o and 1.8o ± 1.7o, respectively, with the virtual fluoroscopy approach. Conclusion: Virtual fluoroscopy could improve accuracy of C-arm positioning and save time and radiation dose in the operating room. Such a system could be valuable to training of fluoroscopy technicians as well as intraoperative use in fluoroscopically guided procedures.

  19. Factors associated with reduced radiation exposure, cost, and technical difficulty of inferior vena cava filter placement and retrieval.

    PubMed

    Neill, Matthew; Charles, Hearns W; Pflager, Daniel; Deipolyi, Amy R

    2017-01-01

    We sought to delineate factors of inferior vena cava filter placement associated with increased radiation and cost and difficult subsequent retrieval. In total, 299 procedures from August 2013 to December 2014, 252 in a fluoroscopy suite (FS) and 47 in the operating room (OR), were reviewed for radiation exposure, fluoroscopy time, filter type, and angulation. The number of retrieval devices and fluoroscopy time needed for retrieval were assessed. Multiple linear regression assessed the impact of filter type, procedure location, and patient and procedural variables on radiation dose, fluoroscopy time, and filter angulation. Logistic regression assessed the impact of filter angulation, type, and filtration duration on retrieval difficulty. Access site and filter type had no impact on radiation exposure. However, placement in the OR, compared to the FS, entailed more radiation (156.3 vs 71.4 mGy; P = 0.001), fluoroscopy time (6.1 vs 2.8 min; P < 0.001), and filter angulation (4.8° vs 2.6°; P < 0.001). Angulation was primarily dependent on filter type ( P = 0.02), with VenaTech and Denali filters associated with decreased angulation (2.2°, 2.4°) and Option filters associated with greater angulation (4.2°). Filter angulation, but not filter type or filtration duration, predicted cases requiring >1 retrieval device ( P < 0.001) and >30 min fluoroscopy time ( P = 0.02). Cost savings for placement in the FS vs OR were estimated at $444.50 per case. In conclusion, increased radiation and cost were associated with placement in the OR. Filter angulation independently predicted difficult filter retrieval; angulation was determined by filter type. Performing filter placement in the FS using specific filters may reduce radiation and cost while enabling future retrieval.

  20. Rapid fusion of 2D X-ray fluoroscopy with 3D multislice CT for image-guided electrophysiology procedures

    NASA Astrophysics Data System (ADS)

    Zagorchev, Lyubomir; Manzke, Robert; Cury, Ricardo; Reddy, Vivek Y.; Chan, Raymond C.

    2007-03-01

    Interventional cardiac electrophysiology (EP) procedures are typically performed under X-ray fluoroscopy for visualizing catheters and EP devices relative to other highly-attenuating structures such as the thoracic spine and ribs. These projections do not however contain information about soft-tissue anatomy and there is a recognized need for fusion of conventional fluoroscopy with pre-operatively acquired cardiac multislice computed tomography (MSCT) volumes. Rapid 2D-3D integration in this application would allow for real-time visualization of all catheters present within the thorax in relation to the cardiovascular anatomy visible in MSCT. We present a method for rapid fusion of 2D X-ray fluoroscopy with 3DMSCT that can facilitate EP mapping and interventional procedures by reducing the need for intra-operative contrast injections to visualize heart chambers and specialized systems to track catheters within the cardiovascular anatomy. We use hardware-accelerated ray-casting to compute digitally reconstructed radiographs (DRRs) from the MSCT volume and iteratively optimize the rigid-body pose of the volumetric data to maximize the similarity between the MSCT-derived DRR and the intra-operative X-ray projection data.

  1. Initial Investigation of Factors Influencing Radiation Dose to Patients Undergoing Barium-Based Fluoroscopy Procedures in Tanzania.

    PubMed

    Ngaile, J E; Msaki, P K; Kazema, R R; Schreiner, L J

    2017-04-25

    The aim of this study was to investigate the nature and causes of radiation dose imparted to patients undergoing barium-based X-ray fluoroscopy procedures in Tanzania and to compare these doses to those reported in the literature from other regions worldwide. The air kerma area product (KAP) to patient undergoing barium investigations of gastrointestinal tract system was obtained from four consultant hospitals. The KAP was determined using a flat transparent transmission ionization chamber. Mean values of KAP for barium swallow (BS), barium meal (BM) and barium enema (BE) were 2.79, 2.62 and 15.04 Gy cm2, respectively. The mean values of KAP per hospital for the BS, BM and BE procedures varied by factors of up to 7.3, 1.6 and 2.0, respectively. The overall difference between individual patient doses across the four consultant hospitals investigated differed by factors of up to 53, 29.5 and 12 for the BS, BM and BE procedures, respectively. The majority of the mean values of KAP was lower than the reported values for Ghana, Greece, Spain and the UK, while slightly higher than those reported for India. The observed wide variation of KAP values for the same fluoroscopy procedure within and among the hospitals was largely attributed to the dynamic nature of the procedures, the patient characteristics, the skills and experience of personnel, and the different examination protocols employed among hospitals. The observed great variations of procedural protocols and patient doses within and across the hospitals call for the need to standardize examination protocols and optimize barium-based fluoroscopy procedures. © The Author 2016. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  2. Effect of pulsed progressive fluoroscopy on reduction of radiation dose in the cardiac catheterization laboratory

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Holmes, D.R. Jr.; Wondrow, M.A.; Gray, J.E.

    1990-01-01

    The increased application of therapeutic interventional cardiology procedures is associated with increased radiation exposure to physicians, patients and technical personnel. New advances in imaging techniques have the potential for reducing radiation exposure. A progressive scanning video system with a standard vascular phantom has been shown to decrease entrance radiation exposure. The effect of this system on reducing actual radiation exposure to physicians and technicians was assessed from 1984 through 1987. During this time, progressive fluoroscopy was added sequentially to all four adult catheterization laboratories; no changes in shielding procedures were made. During this time, the case load per physician increasedmore » by 63% and the number of percutaneous transluminal coronary angioplasty procedures (a high radiation procedure) increased by 244%. Despite these increases in both case load and higher radiation procedures, the average radiation exposure per physician declined by 37%. During the same time, the radiation exposure for technicians decreased by 35%. Pulsed progressive fluoroscopy is effective for reducing radiation exposure to catheterization laboratory physicians and technical staff.« less

  3. Personalized Feedback on Staff Dose in Fluoroscopy-Guided Interventions: A New Era in Radiation Dose Monitoring.

    PubMed

    Sailer, Anna M; Vergoossen, Laura; Paulis, Leonie; van Zwam, Willem H; Das, Marco; Wildberger, Joachim E; Jeukens, Cécile R L P N

    2017-11-01

    Radiation safety and protection are a key component of fluoroscopy-guided interventions. We hypothesize that providing weekly personal dose feedback will increase radiation awareness and ultimately will lead to optimized behavior. Therefore, we designed and implemented a personalized feedback of procedure and personal doses for medical staff involved in fluoroscopy-guided interventions. Medical staff (physicians and technicians, n = 27) involved in fluoroscopy-guided interventions were equipped with electronic personal dose meters (PDMs). Procedure dose data including the dose area product and effective doses from PDMs were prospectively monitored for each consecutive procedure over an 8-month period (n = 1082). A personalized feedback form was designed displaying for each staff individually the personal dose per procedure, as well as relative and cumulative doses. This study consisted of two phases: (1) 1-5th months: Staff did not receive feedback (n = 701) and (2) 6-8th months: Staff received weekly individual dose feedback (n = 381). An anonymous evaluation was performed on the feedback and occupational dose. Personalized feedback was scored valuable by 76% of the staff and increased radiation dose awareness for 71%. 57 and 52% reported an increased feeling of occupational safety and changing their behavior because of personalized feedback, respectively. For technicians, the normalized dose was significantly lower in the feedback phase compared to the prefeedback phase: [median (IQR) normalized dose (phase 1) 0.12 (0.04-0.50) µSv/Gy cm 2 versus (phase 2) 0.08 (0.02-0.24) µSv/Gy cm 2 , p = 0.002]. Personalized dose feedback increases radiation awareness and safety and can be provided to staff involved in fluoroscopy-guided interventions.

  4. Pros and Cons of 3D Image Fusion in Endovascular Aortic Repair: A Systematic Review and Meta-analysis.

    PubMed

    Goudeketting, Seline R; Heinen, Stefan G H; Ünlü, Çağdaş; van den Heuvel, Daniel A F; de Vries, Jean-Paul P M; van Strijen, Marco J; Sailer, Anna M

    2017-08-01

    To systematically review and meta-analyze the added value of 3-dimensional (3D) image fusion technology in endovascular aortic repair for its potential to reduce contrast media volume, radiation dose, procedure time, and fluoroscopy time. Electronic databases were systematically searched for studies published between January 2010 and March 2016 that included a control group describing 3D fusion imaging in endovascular aortic procedures. Two independent reviewers assessed the methodological quality of the included studies and extracted data on iodinated contrast volume, radiation dose, procedure time, and fluoroscopy time. The contrast use for standard and complex endovascular aortic repairs (fenestrated, branched, and chimney) were pooled using a random-effects model; outcomes are reported as the mean difference with 95% confidence intervals (CIs). Seven studies, 5 retrospective and 2 prospective, involving 921 patients were selected for analysis. The methodological quality of the studies was moderate (median 17, range 15-18). The use of fusion imaging led to an estimated mean reduction in iodinated contrast of 40.1 mL (95% CI 16.4 to 63.7, p=0.002) for standard procedures and a mean 70.7 mL (95% CI 44.8 to 96.6, p<0.001) for complex repairs. Secondary outcome measures were not pooled because of potential bias in nonrandomized data, but radiation doses, procedure times, and fluoroscopy times were lower, although not always significantly, in the fusion group in 6 of the 7 studies. Compared with the control group, 3D fusion imaging is associated with a significant reduction in the volume of contrast employed for standard and complex endovascular aortic procedures, which can be particularly important in patients with renal failure. Radiation doses, procedure times, and fluoroscopy times were reduced when 3D fusion was used.

  5. Technique for CT Fluoroscopy-Guided Lumbar Medial Branch Blocks and Radiofrequency Ablation.

    PubMed

    Amrhein, Timothy J; Joshi, Anand B; Kranz, Peter G

    2016-09-01

    The purpose of this study is to describe the procedure for CT fluoroscopy-guided lumbar medial branch blocks and facet radiofrequency ablation. CT fluoroscopic guidance allows more-precise needle tip positioning and is an alternative method for performing medial branch blocks and facet radiofrequency ablation.

  6. Monte Carlo based estimation of organ and effective doses to patients undergoing hysterosalpingography and retrograde urethrography fluoroscopy procedures

    NASA Astrophysics Data System (ADS)

    Ngaile, J. E.; Msaki, P. K.; Kazema, R. R.

    2018-04-01

    Contrast investigations of hysterosalpingography (HSG) and retrograde urethrography (RUG) fluoroscopy procedures remain the dominant diagnostic tools for the investigation of infertility in females and urethral strictures in males, respectively, owing to the scarcity and high cost of services of alternative diagnostic technologies. In light of the radiological risks associated with contrast based investigations of the genitourinary tract systems, there is a need to assess the magnitude of radiation burden imparted to patients undergoing HSG and RUG fluoroscopy procedures in Tanzania. The air kerma area product (KAP), fluoroscopy time, number of images, organ dose and effective dose to patients undergoing HSG and RUG procedures were obtained from four hospitals. The KAP was measured using a flat transmission ionization chamber, while the organ and effective doses were estimated using the knowledge of the patient characteristics, patient related exposure parameters, geometry of examination, KAP and Monte Carlo calculations (PCXMC). The median values of KAP for the HSG and RUG were 2.2 Gy cm2 and 3.3 Gy cm2, respectively. The median organ doses in the present study for the ovaries, urinary bladder and uterus for the HSG procedures, were 1.0 mGy, 4.0 mGy and 1.6 mGy, respectively, while for urinary bladder and testes of the RUG were 3.4 mGy and 5.9 mGy, respectively. The median values of effective doses for the HSG and RUG procedures were 0.65 mSv and 0.59 mSv, respectively. The median values of effective dose per hospital for the HSG and RUG procedures had a range of 1.6-2.8 mSv and 1.9-5.6 mSv, respectively, while the overall differences between individual effective doses across the four hospitals varied by factors of up to 22.0 and 46.7, respectively for the HSG and RUG procedures. The proposed diagnostic reference levels (DRLs) for the HSG and RUG were for KAP 2.8 Gy cm2 and 3.9 Gy cm2, for fluoroscopy time 0.8 min and 0.9 min, and for number of images 5 and 4, respectively. The suggested DRLs for the HSG and RUG procedures may be used by the radiology departments in Tanzania for management of attained dose levels until the national DRLs are established.

  7. Electromagnetic navigation versus fluoroscopy in aortic endovascular procedures: a phantom study.

    PubMed

    Tystad Lund, Kjetil; Tangen, Geir Arne; Manstad-Hulaas, Frode

    2017-01-01

    To explore the possible benefits of electromagnetic (EM) navigation versus conventional fluoroscopy during abdominal aortic endovascular procedures. The study was performed on a phantom representing the abdominal aorta. Intraoperative cone beam computed tomography (CBCT) of the phantom was acquired and merged with a preoperative multidetector CT (MDCT). The CBCT was performed with a reference plate fixed to the phantom that, after merging the CBCT with the MDCT, facilitated registration of the MDCT volume with the EM space. An EM field generator was stationed near the phantom. Navigation software was used to display EM-tracked instruments within the 3D image volume. Fluoroscopy was performed using a C-arm system. Five operators performed a series of renal artery cannulations using modified instruments, alternatingly using fluoroscopy or EM navigation as the sole guidance method. Cannulation durations and associated radiation dosages were noted along with the number of cannulations complicated by loss of guidewire insertion. A total of 120 cannulations were performed. The median cannulation durations were 41.5 and 34.5 s for the fluoroscopy- and EM-guided cannulations, respectively. No significant difference in cannulation duration was found between the two modalities (p = 0.736). Only EM navigation showed a significant reduction in cannulation duration in the latter half of its cannulation series compared with the first half (p = 0.004). The median dose area product for fluoroscopy was 0.0836 [Formula: see text]. EM-guided cannulations required a one-time CBCT dosage of 3.0278 [Formula: see text]. Three EM-guided and zero fluoroscopy-guided cannulations experienced loss of guidewire insertion. Our findings indicate that EM navigation is not inferior to fluoroscopy in terms of the ability to guide endovascular interventions. Its utilization may be of particular interest in complex interventions where adequate visualization or minimal use of contrast agents is critical. In vivo studies featuring an optimized implementation of EM navigation should be conducted.

  8. Fluoroscopy-assisted vs fluoroless endoscopic ultrasound-guided transmural drainage of pancreatic fluid collections: A comparative study.

    PubMed

    Consiglieri, Claudia F; Gornals, Joan B; Busquets, Juli; Peláez, Nuria; Secanella, Lluis; De-La-Hera, Meritxell; Sanzol, Resurrección; Fabregat, Joan; Castellote, José

    2018-01-01

    The need for fluoroscopy guidance in patients undergoing endoscopic ultrasound-guided transmural drainage (EUS-TMD) of peripancreatic fluid collections (PFCs) remains unclear. The aim of this study was to compare general outcomes of EUS-TMD of PFCs under fluoroscopy (F) vs fluoroless (FL). This is a comparative study with a retrospective analysis of a prospective and consecutive inclusion database at a tertiary centre, from 2009 to 2015. All patients were symptomatic pseudocyst (PSC) and walled-off pancreatic necrosis (WON). Two groups were assigned depending on availability of fluoroscopy. The groups were heterogeneous in terms of their demographic characteristics, PFCs and procedure. The main outcome measures included technical and clinical success, incidences, adverse events (AEs), and follow-up. Fifty EUS-TMD of PFCs from 86 EUS-guided drainages were included during the study period. Group F included 26 procedures, PSC 69.2%, WON 30.8%, metal stents 61.5% (46.1% lumen-apposing stent) and plastic stents 38.5%. Group FL included 24 procedures, PSC 37.5%, WON 62.5%, and metal stents 95.8% (lumen-apposing stents). Technical success was 100% in both groups, and clinical success was similar (F 88.5%, FL 87.5%). Technical incidences and intra-procedure AEs were only described in group F (7.6% and 11.5%, respectively) and none in group FL. Procedure time was less in group FL (8min, p=0.0341). Fluoroless in the EUS-TMD of PFCs does not involve more technical incidences or intra-procedure AEs. Technical and clinical success was similar in the two groups. Copyright © 2017 Elsevier España, S.L.U. All rights reserved.

  9. Predicting the difficulty of a lead extraction procedure: the LED index.

    PubMed

    Bontempi, Luca; Vassanelli, Francesca; Cerini, Manuel; D'Aloia, Antonio; Vizzardi, Enrico; Gargaro, Alessio; Chiusso, Francesco; Mamedouv, Rashad; Lipari, Alessandro; Curnis, Antonio

    2014-08-01

    According to recent surveys, many sites performing permanent lead extractions do not meet the minimum prerequisites concerning personnel training, procedures' volume, or facility requirements. The current Heart Rhythm Society consensus on lead extractions suggests that patients should be referred to more experienced sites when a better outcome could be achieved. The purpose of this study was to develop a score aimed at predicting the difficulty of a lead extraction procedure through the analysis of a high-volume center database. This score could help to discriminate patients who should be sent to a referral site. A total of 889 permanent leads were extracted from 469 patients. All procedures were performed from January 2009 to May 2012 by two expert electrophysiologists, at the University Hospital of Brescia. Factors influencing the difficulty of a procedure were assessed using a univariate and a multivariate logistic regression model. The fluoroscopy time of the procedure was taken as an index of difficulty. A Lead Extraction Difficulty (LED) score was defined, considering the strongest predictors. Overall, 873 of 889 (98.2%) leads were completely removed. Major complications were reported in one patient (0.2%) who manifested cardiac tamponade. Minor complications occurred in six (1.3%) patients. No deaths occurred. Median fluoroscopic time was 8.7 min (3.3-17.3). A procedure was classified as difficult when fluoroscopy time was more than 31.2 min [90th percentile (PCTL)].At a univariate analysis, the number of extracted leads and years from implant were significantly associated with an increased risk of fluoroscopy time above 90th PCTL [odds ratio (OR) 1.51, 95% confidence interval (CI) 1.08-2.11, P = 0.01; and OR 1.19, 95% CI 1.12-1.25, P < 0.001, respectively). After adjusting for patient age and sex, and combining with other covariates potentially influencing the extraction procedure, a multivariate analysis confirmed a 71% increased risk of fluoroscopy time above 90th PCTL for each additional lead extracted (OR 1.71, 95% CI 1.06-2.77, P = 0.028) and a 23% increased risk for each year of lead age (OR 1.23, 95% CI 1.15-1.31, P < 0.001). Further nonindependent factors increasing the risk were the presence of active fixation leads and dual-coil implantable cardiac defibrillator leads. Conversely, vegetations significantly favored lead extraction.The LED score was defined as: number of extracted leads within a procedure + lead age (years from implant) + 1 if dual-coil - 1 if vegetation. The LED score independently predicted complex procedure (with fluoroscopic time >90th PCTL) both at univariate and multivariate analysis. A receiver-operating characteristic analysis showed an area under the curve of 0.81. A LED score greater than 10 could predict fluoroscopy time above 90th PCTL with a sensitivity of 78.3% and a specificity of 76.7%. The LED score is easy to compute and potentially predicts fluoroscopy time above 90th PCTL with a relatively high accuracy.

  10. Biplanar x-ray fluoroscopy for sacroiliac joint fusion.

    PubMed

    Vanaclocha-Vanaclocha, Vicente; Verdú-López, Francisco; Sáiz-Sapena, Nieves; Herrera, Juan Manuel; Rivera-Paz, Marlon

    2016-07-01

    Chronic pain originating from the sacroiliac joint (SI) can cause severe dysfunction. Although many patients respond to conservative management with NSAIDs, some do need further treatment in the form of SI joint fusion (SIJF). To achieve safe and successful SIJF, intraoperative x-ray fluoroscopy is mandatory to avoid serious damages to nearby vascular and neural structures. Each step of the procedure has to be confirmed by anteroposterior (AP) and lateral projections. With a single-arm x-ray, the arch has to be moved back and forth for the AP and lateral projections, and this lengthens the procedure. To achieve the same results in less time, the authors introduced simultaneous biplanar fluoroscopy with 2 x-ray arches. After the patient is positioned prone with the legs spread apart in the so-called Da Vinci position, one x-ray arch for the lateral projection is placed at a right angle to the patient, and a second x-ray machine is placed with its arch between the legs of the patient. This allows simultaneous AP and lateral x-ray projections and, in the authors' hands, markedly speeds up the procedure. Biplanar fluoroscopy allows excellent AP and lateral projections to be made quickly at any time during the surgical procedure. This is particularly useful in cases of bilateral SI joint fusion if both sides are done at the same time. The video can be found here: https://youtu.be/TX5gz8c765M .

  11. Fluoroscopy-guided Sacroiliac Joint Steroid Injection for Low Back Pain in a Patient with Osteogenesis Imperfecta.

    PubMed

    Dawson, P U; Rose, R E; Wade, N A

    2015-09-01

    Osteogenesis imperfecta, also known as 'brittle bone disease', is a genetic connective tissue disease. It is characterized by bone fragility and osteopenia (low bone density). In this case, a 57-year old female presented to the University Hospital of the West Indies (UHWI), Physical Medicine and Rehabilitation Clinic with left low back pain rated 6/10 on the numeric rating scale (NRS). Clinically, the patient had sacroiliac joint mediated pain although X-rays did not show the sacroiliac joint changes. Fluoroscopy-guided left sacroiliac joint steroid injection was done. Numeric rating scale and Oswestry Disability Index (ODI) questionnaire were used to evaluate outcome. This was completed at baseline, one week follow-up and at eight weeks post fluoroscopy-guided sacroiliac joint steroid injection. Numeric rating scale improved from 6/10 before the procedure to 0/10 post procedure, and ODI questionnaire score improved from a moderate disability score of 40% to a minimal disability score of 13%. Up to eight weeks, the NRS was 0/10 and ODI remained at minimal disability of 15%. Fluoroscopy-guided sacroiliac joint injection is a known diagnostic and treatment method for sacroiliac joint mediated pain. To our knowledge, this is the first case published on the use of fluoroscopy-guided sacroiliac joint steroid injection in the treatment of sacroiliac joint mediated low back pain in a patient with osteogenesis imperfecta.

  12. Accuracy assessment of fluoroscopy-transesophageal echocardiography registration

    NASA Astrophysics Data System (ADS)

    Lang, Pencilla; Seslija, Petar; Bainbridge, Daniel; Guiraudon, Gerard M.; Jones, Doug L.; Chu, Michael W.; Holdsworth, David W.; Peters, Terry M.

    2011-03-01

    This study assesses the accuracy of a new transesophageal (TEE) ultrasound (US) fluoroscopy registration technique designed to guide percutaneous aortic valve replacement. In this minimally invasive procedure, a valve is inserted into the aortic annulus via a catheter. Navigation and positioning of the valve is guided primarily by intra-operative fluoroscopy. Poor anatomical visualization of the aortic root region can result in incorrect positioning, leading to heart valve embolization, obstruction of the coronary ostia and acute kidney injury. The use of TEE US images to augment intra-operative fluoroscopy provides significant improvements to image-guidance. Registration is achieved using an image-based TEE probe tracking technique and US calibration. TEE probe tracking is accomplished using a single-perspective pose estimation algorithm. Pose estimation from a single image allows registration to be achieved using only images collected in standard OR workflow. Accuracy of this registration technique is assessed using three models: a point target phantom, a cadaveric porcine heart with implanted fiducials, and in-vivo porcine images. Results demonstrate that registration can be achieved with an RMS error of less than 1.5mm, which is within the clinical accuracy requirements of 5mm. US-fluoroscopy registration based on single-perspective pose estimation demonstrates promise as a method for providing guidance to percutaneous aortic valve replacement procedures. Future work will focus on real-time implementation and a visualization system that can be used in the operating room.

  13. CT Fluoroscopy Shielding: Decreases in Scattered Radiation for the Patient and Operator

    PubMed Central

    Neeman, Ziv; Dromi, Sergio A.; Sarin, Shawn; Wood, Bradford J.

    2008-01-01

    PURPOSE High-radiation exposure occurs during computed tomographic (CT) fluoroscopy. Patient and operator doses during thoracic and abdominal interventional procedures were studied in the present experiment, and a novel shielding device to reduce exposure to the patient and operator was evaluated. MATERIALS AND METHODS With a 16-slice CT scanner in CT fluoroscopy mode (120 kVp, 30 mA), surface dosimetry was performed on adult and pediatric phantoms. The shielding was composed of tungsten antimony in the form of a lightweight polymer sheet. Doses to the patient were measured with and without shielding for thoracic and abdominal procedures. Doses to the operator were recorded with and without phantom, gantry, and table shielding in place. Double-layer lead-free gloves were used by the operator during the procedures. RESULTS Tungsten antimony shielding adjacent to the scan plane resulted in a maximum dose reduction of 92.3% to the patient. Maximum 85.6%, 93.3%, and 85.1% dose reductions were observed for the operator’s torso, gonads, and hands, respectively. The use of double-layer lead-free gloves resulted in a maximum radiation dose reduction of 97%. CONCLUSIONS Methods to reduce exposure during CT fluoroscopy are effective and should be searched for. Significant reduction in radiation doses to the patient and operator can be accomplished with tungsten antimony shielding. PMID:17185699

  14. Comparative ergonomic workflow and user experience analysis of MRI versus fluoroscopy-guided vascular interventions: an iliac angioplasty exemplar case study.

    PubMed

    Fernández-Gutiérrez, Fabiola; Martínez, Santiago; Rube, Martin A; Cox, Benjamin F; Fatahi, Mahsa; Scott-Brown, Kenneth C; Houston, J Graeme; McLeod, Helen; White, Richard D; French, Karen; Gueorguieva, Mariana; Immel, Erwin; Melzer, Andreas

    2015-10-01

    A methodological framework is introduced to assess and compare a conventional fluoroscopy protocol for peripheral angioplasty with a new magnetic resonant imaging (MRI)-guided protocol. Different scenarios were considered during interventions on a perfused arterial phantom with regard to time-based and cognitive task analysis, user experience and ergonomics. Three clinicians with different expertise performed a total of 43 simulated common iliac angioplasties (9 fluoroscopic, 34 MRI-guided) in two blocks of sessions. Six different configurations for MRI guidance were tested in the first block. Four of them were evaluated in the second block and compared to the fluoroscopy protocol. Relevant stages' durations were collected, and interventions were audio-visually recorded from different perspectives. A cued retrospective protocol analysis (CRPA) was undertaken, including personal interviews. In addition, ergonomic constraints in the MRI suite were evaluated. Significant differences were found when comparing the performance between MRI configurations versus fluoroscopy. Two configurations [with times of 8.56 (0.64) and 9.48 (1.13) min] led to reduce procedure time for MRI guidance, comparable to fluoroscopy [8.49 (0.75) min]. The CRPA pointed out the main influential factors for clinical procedure performance. The ergonomic analysis quantified musculoskeletal risks for interventional radiologists when utilising MRI. Several alternatives were suggested to prevent potential low-back injuries. This work presents a step towards the implementation of efficient operational protocols for MRI-guided procedures based on an integral and multidisciplinary framework, applicable to the assessment of current vascular protocols. The use of first-user perspective raises the possibility of establishing new forms of clinical training and education.

  15. Training for percutaneous renal access on a virtual reality simulator.

    PubMed

    Zhang, Yi; Yu, Cheng-fan; Liu, Jin-shun; Wang, Gang; Zhu, He; Na, Yan-qun

    2013-01-01

    The need to develop new methods of surgical training combined with advances in computing has led to the development of virtual reality surgical simulators. The PERC Mentor(TM) is designed to train the user in percutaneous renal collecting system access puncture. This study aimed to validate the use of this kind of simulator, in percutaneous renal access training. Twenty-one urologists were enrolled as trainees to learn a fluoroscopy-guided percutaneous renal accessing technique. An assigned percutaneous renal access procedure was immediately performed on the PERC Mentor(TM) after watching instruction video and an analog operation. Objective parameters were recorded by the simulator and subjective global rating scale (GRS) score were determined. Simulation training followed and consisted of 2 hours daily training sessions for 2 consecutive days. Twenty-four hours after the training session, trainees were evaluated performing the same procedure. The post-training evaluation was compared to the evaluation of the initial attempt. During the initial attempt, none of the trainees could complete the appointed procedure due to the lack of experience in fluoroscopy-guided percutaneous renal access. After the short-term training, all trainees were able to independently complete the procedure. Of the 21 trainees, 10 had primitive experience in ultrasound-guided percutaneous nephrolithotomy. Trainees were thus categorized into the group of primitive experience and inexperience. The total operating time and amount of contrast material used were significantly lower in the group of primitive experience versus the inexperience group (P = 0.03 and 0.02, respectively). The training on the virtual reality simulator, PERC Mentor(TM), can help trainees with no previous experience of fluoroscopy-guided percutaneous renal access to complete the virtual manipulation of the procedure independently. This virtual reality simulator may become an important training and evaluation tool in teaching fluoroscopy-guided percutaneous renal access.

  16. 42 CFR 411.351 - Definitions.

    Code of Federal Regulations, 2012 CFR

    2012-10-01

    ... professional and technical components of any diagnostic test or procedure using x-rays, ultrasound...-ray, fluoroscopy, or ultrasound procedures that require the insertion of a needle, catheter, tube, or...

  17. 42 CFR 411.351 - Definitions.

    Code of Federal Regulations, 2013 CFR

    2013-10-01

    ... professional and technical components of any diagnostic test or procedure using x-rays, ultrasound...-ray, fluoroscopy, or ultrasound procedures that require the insertion of a needle, catheter, tube, or...

  18. 42 CFR 411.351 - Definitions.

    Code of Federal Regulations, 2014 CFR

    2014-10-01

    ... professional and technical components of any diagnostic test or procedure using x-rays, ultrasound...-ray, fluoroscopy, or ultrasound procedures that require the insertion of a needle, catheter, tube, or...

  19. Is the 10th and 11th Intercostal Space a Safe Approach for Percutaneous Nephrostomy and Nephrolithotomy?

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Muzrakchi, Ahmed Al; Szmigielski, W., E-mail: wojszmi@qatar.net.qa; Omar, Ahmed J.S.

    2004-09-15

    The aim of this study was to determine the rate of complications in percutaneous nephrostomy (PCN) and nephrolithotomy (PCNL) performed through the 11th and 10th intercostal spaces using our monitoring technique and to discuss the safety of the procedure. Out of 398 PCNs and PCNLs carried out during a 3-year period, 56 patients had 57 such procedures performed using an intercostal approach. The 11th intercostal route was used in 42 and the 10th in 15 cases. One patient had two separate nephrostomies performed through the 10th and 11th intercostal spaces. The technique utilizes bi-planar fluoroscopy with a combination of amore » conventional angiographic machine to provide anterior-posterior fluoroscopy and a C-arm mobile fluoroscopy machine to give a lateral view, displayed on two separate monitors. None of the patients had clinically significant thoracic or abdominal complications. Two patients had minor chest complications. Only one developed changes (plate atelectasis, elevation of the hemi-diaphragm) directly related to the nephrostomy (2%). The second patient had bilateral plate atelectasis and unilateral congestive lung changes after PCNL. These changes were not necessarily related to the procedure but rather to general anesthesia during nephrolithotomy. The authors consider PCN or PCNL through the intercostal approach a safe procedure with a negligible complication rate, provided that it is performed under bi-planar fluoroscopy, which allows determination of the skin entry point just below the level of pleural reflection and provides three-dimensional monitoring of advancement of the puncturing needle toward the target entry point.« less

  20. CT fluoroscopy-guided robotically-assisted lung biopsy

    NASA Astrophysics Data System (ADS)

    Xu, Sheng; Fichtinger, Gabor; Taylor, Russell H.; Banovac, Filip; Cleary, Kevin

    2006-03-01

    Lung biopsy is a common interventional radiology procedure. One of the difficulties in performing the lung biopsy is that lesions move with respiration. This paper presents a new robotically assisted lung biopsy system for CT fluoroscopy that can automatically compensate for the respiratory motion during the intervention. The system consists of a needle placement robot to hold the needle on the CT scan plane, a radiolucent Z-frame for registration of the CT and robot coordinate systems, and a frame grabber to obtain the CT fluoroscopy image in real-time. The CT fluoroscopy images are used to noninvasively track the motion of a pulmonary lesion in real-time. The position of the lesion in the images is automatically determined by the image processing software and the motion of the robot is controlled to compensate for the lesion motion. The system was validated under CT fluoroscopy using a respiratory motion simulator. A swine study was also done to show the feasibility of the technique in a respiring animal.

  1. Fluoroscopic exposure in modern spinal surgery.

    PubMed

    Fransen, Patrick

    2011-06-01

    The widespread use of minimally invasive and other spinal procedures raises concern about the peroperative radiation exposure to surgeon and patient. The authors noted the fluoroscopy time and the radiation dose, as read from the image amplifier, in 95 spinal procedures. The results of this prospective study varied widely between different operations. Percutaneous surgery was associated with more exposure than open surgery. For instance, the average radiation dose per pedicle screw was 3.2 times higher with percutaneous insertion than with an open approach. Therefore, efforts to reduce fluoroscopy time and radiation exposure should be made when using minimally invasive percutaneous surgical techniques. Preventive measures for the surgeon, such as lead aprons and gloves, thyroid shields, radioprotective glasses and staying away from the beam are recommended. Still from the surgeon's view-point, source inferior positioning of the image amplifier is indicated for the AP view, as well as monitoring of the radiation exposure. Finally, the difference in fluoroscopy time and radiation exposure between surgeons for the same procedure stresses the fact that peroperative radiation may be reduced by simple awareness and by training.

  2. Is There a Relationship Between Body Mass Index and Fluoroscopy Time During Sacroiliac Joint Injection? A Multicenter Cohort Study.

    PubMed

    McCormick, Zachary L; Cushman, Daniel; Lee, David T; Scholten, Paul; Chu, Samuel K; Babu, Ashwin N; Caldwell, Mary; Ziegler, Craig; Ashraf, Humaira; Sundar, Bindu; Clark, Ryan; Gross, Claire; Cara, Jeffrey; McCormick, Kristen; Ross, Brendon; Smith, Clark C; Press, Joel; Smuck, Matthew; Walega, David R

    2016-07-01

    To determine the relationship between BMI and fluoroscopy time during intra-articular sacroiliac joint (SIJ) injections performed for a pain indication. Multicenter retrospective cohort study. Three academic, outpatient pain treatment centers. Patients who underwent fluoroscopy guided SIJ injection with encounter data regarding fluoroscopy time during the procedure and body mass index (BMI). Median and 25-75% Interquartile Range (IQR) fluoroscopy time. 459 SIJ injections (350 patients) were included in this study. Patients had a median age of 57 (IQR 44, 70) years, and 72% were female. The median BMI in the normal weight, overweight, and obese groups were 23 (IQR 21, 24), 27 (IQR 26, 29), and 35 (IQR 32, 40), respectively. There was no significant difference in the median fluoroscopy time recorded between these BMI classes (p = 0.45). First-time SIJ injection (p = 0.53), bilateral injection (p = 0.30), trainee involvement (p = 0.47), and new trainee involvement (trainee participation during the first 2 months of the academic year) (p = 0.85) were not associated with increased fluoroscopy time for any of the three BMI categories. Fluoroscopy time during sacroiliac joint injection is not increased in patients who are overweight or obese, regardless of whether a first-time sacroiliac joint injection was performed, bilateral injections were performed, a trainee was involved, or a new trainee was involved. © 2015 American Academy of Pain Medicine. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.

  3. Detection of electrophysiology catheters in noisy fluoroscopy images.

    PubMed

    Franken, Erik; Rongen, Peter; van Almsick, Markus; ter Haar Romeny, Bart

    2006-01-01

    Cardiac catheter ablation is a minimally invasive medical procedure to treat patients with heart rhythm disorders. It is useful to know the positions of the catheters and electrodes during the intervention, e.g. for the automatization of cardiac mapping. Our goal is therefore to develop a robust image analysis method that can detect the catheters in X-ray fluoroscopy images. Our method uses steerable tensor voting in combination with a catheter-specific multi-step extraction algorithm. The evaluation on clinical fluoroscopy images shows that especially the extraction of the catheter tip is successful and that the use of tensor voting accounts for a large increase in performance.

  4. Impact of operator experience and training strategy on procedural outcomes with leadless pacing: Insights from the Micra Transcatheter Pacing Study.

    PubMed

    El-Chami, Mikhael; Kowal, Robert C; Soejima, Kyoko; Ritter, Philippe; Duray, Gabor Z; Neuzil, Petr; Mont, Lluis; Kypta, Alexander; Sagi, Venkata; Hudnall, John Harrison; Stromberg, Kurt; Reynolds, Dwight

    2017-07-01

    Leadless pacemaker systems have been designed to avoid the need for a pocket and transvenous lead. However, delivery of this therapy requires a new catheter-based procedure. This study evaluates the role of operator experience and different training strategies on procedural outcomes. A total of 726 patients underwent implant attempt with the Micra transcatheter pacing system (TPS; Medtronic, Minneapolis, MN, USA) by 94 operators trained in a teaching laboratory using a simulator, cadaver, and large animal models (lab training) or locally at the hospital with simulator/demo model and proctorship (hospital training). Procedure success, procedure duration, fluoroscopy time, and safety outcomes were compared between training methods and experience (implant case number). The Micra TPS procedure was successful in 99.2% of attempts and did not differ between the 55 operators trained in the lab setting and the 39 operators trained locally at the hospital (P = 0.189). Implant case number was also not a determinant of procedural success (P = 0.456). Each operator performed between one and 55 procedures. Procedure time and fluoroscopy duration decreased by 2.0% (P = 0.002) and 3.2% (P < 0.001) compared to the previous case. Major complication rate and pericardial effusion rate were not associated with case number (P = 0.755 and P = 0.620, respectively). There were no differences in the safety outcomes by training method. Among a large group of operators, implantation success was high regardless of experience. While procedure duration and fluoroscopy times decreased with implant number, complications were low and not associated with case number. Procedure and safety outcomes were similar between distinct training methodologies. © 2017 Wiley Periodicals, Inc.

  5. "Turn laterally to the left!". The need for uniform C-arm communication terminology during orthopaedic trauma surgery.

    PubMed

    De Muinck Keizer, R-J; Klei, D S; Van Koperen, P J; Van Dijk, C N; Goslings, J C

    2017-03-01

    To avoid disturbed teamwork, unnecessary radiation exposure, and procedural delays, we designed and tested a uniform communication language for use in fluoroscopy-assisted surgical procedures. Input of surgeons and radiographers was used to create a set of commands. The potential benefit of this terminology was explored in an experimental setting. There was a tremendous diversity in the currently used terminology. Use of the newly designed terminology showed a reduction of procedural time and amount of images needed. Our first standardized Dutch language terminology can reduce total fluoroscopy time, number of images acquired, and potentially radiation exposure. For Dutch speaking colleagues, the developed terminology is freely available for use in their OR.

  6. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jones, A. Kyle, E-mail: kyle.jones@mdanderson.org

    Purpose: To evaluate the sensitivity of the diagnostic radiological index of protection (DRIP), used to quantify the protective value of radioprotective garments, to procedural factors in fluoroscopy in an effort to determine an appropriate set of scatter-mimicking primary beams to be used in measuring the DRIP. Methods: Monte Carlo simulations were performed to determine the shape of the scattered x-ray spectra incident on the operator in different clinical fluoroscopy scenarios, including interventional radiology and interventional cardiology (IC). Two clinical simulations studied the sensitivity of the scattered spectrum to gantry angle and patient size, while technical factors were varied according tomore » measured automatic dose rate control (ADRC) data. Factorial simulations studied the sensitivity of the scattered spectrum to gantry angle, field of view, patient size, and beam quality for constant technical factors. Average energy (E{sub avg}) was the figure of merit used to condense fluence in each energy bin to a single numerical index. Results: Beam quality had the strongest influence on the scattered spectrum in fluoroscopy. Many procedural factors affect the scattered spectrum indirectly through their effect on primary beam quality through ADRC, e.g., gantry angle and patient size. Lateral C-arm rotation, common in IC, increased the energy of the scattered spectrum, regardless of the direction of rotation. The effect of patient size on scattered radiation depended on ADRC characteristics, patient size, and procedure type. Conclusions: The scattered spectrum striking the operator in fluoroscopy is most strongly influenced by primary beam quality, particularly kV. Use cases for protective garments should be classified by typical procedural primary beam qualities, which are governed by the ADRC according to the impacts of patient size, anatomical location, and gantry angle.« less

  7. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jones, A; Pasciak, A; Wagner, L

    Purpose: To evaluate the sensitivity of the Diagnostic Radiological Index of Protection (DRIP) to procedural factors in fluoroscopy in an effort to determine an appropriate set of scatter-mimicking primary beams (SMPB) to be used in measuring the DRIP. Methods: A series of clinical and factorial Monte Carlo simulations were conducted to determine the shape of the scattered X-ray spectra incident on the operator in different clinical fluoroscopy scenarios. Two clinical evaluations studied the sensitivity of the scattered spectrum to gantry angle and patient size while technical factors were varied according to measured automatic dose rate control (ADRC) data. Factorial evaluationsmore » studied the sensitivity of the scattered spectrum to gantry angle, field of view, patient size and beam quality for constant technical factors. Average energy was the figure of merit used to condense fluence in each energy bin to a single numerical index. Results: Beam quality had the strongest influence on the scattered spectrum in fluoroscopy. Many procedural factors affected the scattered spectrum indirectly through their effects on primary beam quality through ADRC, e.g., gantry angle and patient size. Lateral C-arm rotation, common in interventional cardiology, increased the energy of the scattered spectrum, regardless of the direction of rotation. The effect of patient size on scattered radiation depended on ADRC characteristics, patient size, and procedure type. Conclusion: The scattered spectrum striking the operator in fluoroscopy, and therefore the DRIP, is most strongly influenced by primary beam quality, particularly kV. Use cases for protective garments should be classified by typical procedural primary beam qualities, which are governed by the ADRC according to the impacts of patient size, anatomical location, and gantry angle. These results will help determine an appropriate set of SMPB to be used for measuring the DRIP.« less

  8. Predictors of thrombotic complications and mass effect exacerbation after pipeline embolization: The significance of adenosine diphosphate inhibition, fluoroscopy time, and aneurysm size.

    PubMed

    Raychev, Radoslav; Tateshima, Satoshi; Vinuela, Fernando; Sayre, Jim; Jahan, Reza; Gonzalez, Nestor; Szeder, Viktor; Duckwiler, Gary

    2016-02-01

    The mechanisms leading to delayed rupture, distal emboli and intraparenchymal hemorrhage in relation to pipeline embolization device (PED) placement remain debatable and poorly understood. The aim of this study was to identify clinical and procedural predictors of these perioperative complications. We conducted a retrospective review of consecutive patients who underwent PED placement. We utilized a non-commercial platelet aggregation method measuring adenosine diphosphate (ADP)% inhibition for evaluation of clopidogrel response. To our knowledge, this is the first study to test ADP in neurovascular procedures. Multivariable regression analysis was used to identify the strongest predictor of three separate outcomes: (1) thrombotic complications, (2) hemorrhagic complications, and (3) aneurysm mass effect exacerbation Permanent complication-related morbidity and mortality at 3 months was 6% (3/48). No specific predictors of hemorrhagic complications were identified. In the univariate analysis, the strongest predictors of thrombotic complications were: ADP% inhibition<49 (p=0.01), aneurysm size (p=0.04) and fluoroscopy time (p=0.002). In the final multivariate analysis, among all baseline variables, fluoroscopy time exceeding 52 min was the only factor associated with thrombotic complications (p=0.007). Aneurysm size≥18 mm was the single predictor of mass effect exacerbation (p=0.039). Procedural complexity, reflected by fluoroscopy time, is the strongest predictor of thrombotic complications in this study. ADP% inhibition is a reliable method of testing clopidogrel response in neurovascular procedures and values of <50% may predict thrombotic complications. Interval mass effect exacerbation after PED placement may be anticipated in large aneurysms exceeding 18 mm. © The Author(s) 2015.

  9. Guidewire navigation in coronary artery stenoses using a novel magnetic navigation system: first clinical experience.

    PubMed

    Tsuchida, Keiichi; García-García, Héctor M; van der Giessen, Willem J; McFadden, Eugène P; van der Ent, Martin; Sianos, Georgios; Meulenbrug, Hans; Ong, Andrew T L; Serruys, Patrick W

    2006-03-01

    The objective of this study was to investigate the efficacy of guidewire navigation across coronary artery stenoses using magnetic navigation system (MNS) versus conventional navigation. The MNS is a novel option to facilitate access to target lesions, particularly in tortuous vessels. In an experimental study using a challenging vessel phantom, magnetic-navigated guidewire passage has been reported to reduce fluoroscopy and procedure time significantly. Both magnetic and manual guidewire navigation were attempted in 21 consecutive diseased coronary arteries. The study endpoint was defined as an intraluminal wire position distal to the stenosis. Procedural success was defined as successful guidewire passage without procedural events. Procedure time, amount of contrast, fluoroscopy time, and radiation dose/area product (DAP) were evaluated. There were no procedural events related to either guidewire. Although the lesions attempted had relatively simple and straightforward characteristics, significantly shorter procedure and fluoroscopy time were observed for manual guidewire navigation compared to MNS (median, 40 vs. 120 sec, P=0.001; 38 vs. 105 sec, P=0.001, respectively). Contrast amount and DAP were higher in MNS than in conventional method (median, 13 vs. 9 ml, P=0.018; 215 vs. 73 Gym2, P=0.002, respectively). The magnetic wire did not cross in two vessels. Guidewire navigation using MNS presented a novel, safe, and feasible approach to address coronary artery lesions. Clinical studies are needed to evaluate the potential benefit of the MNS in more complex coronary lesions and tortuous anatomy. Copyright (c) 2006 Wiley-Liss, Inc.

  10. Preliminary study of rib articulated model based on dynamic fluoroscopy images

    NASA Astrophysics Data System (ADS)

    Villard, Pierre-Frederic; Escamilla, Pierre; Kerrien, Erwan; Gorges, Sebastien; Trousset, Yves; Berger, Marie-Odile

    2014-03-01

    We present in this paper a preliminary study of rib motion tracking during Interventional Radiology (IR) fluoroscopy guided procedures. It consists in providing a physician with moving rib three-dimensional (3D) models projected in the fluoroscopy plane during a treatment. The strategy is to help to quickly recognize the target and the no-go areas i.e. the tumor and the organs to avoid. The method consists in i) elaborating a kinematic model of each rib from a preoperative computerized tomography (CT) scan, ii) processing the on-line fluoroscopy image and iii) optimizing the parameters of the kinematic law such as the transformed 3D rib projected on the medical image plane fit well with the previously processed image. The results show a visually good rib tracking that has been quantitatively validated by showing a periodic motion as well as a good synchronism between ribs.

  11. Fluoroscopy guided percutaneous renal access in prone position

    PubMed Central

    Sharma, Gyanendra R; Maheshwari, Pankaj N; Sharma, Anshu G; Maheshwari, Reeta P; Heda, Ritwik S; Maheshwari, Sakshi P

    2015-01-01

    Percutaneous nephrolithotomy is a very commonly done procedure for management of renal calculus disease. Establishing a good access is the first and probably the most crucial step of this procedure. A proper access is the gateway to success. However, this crucial step has the steepest learning curve for, in a fluoroscopy guided access, it involves visualizing a three dimensional anatomy on a two dimensional fluoroscopy screen. This review describes the anatomical basis of the renal access. It provides a literature review of all aspects of percutaneous renal access along with the advances that have taken place in this field over the years. The article describes a technique to determine the site of skin puncture, the angle and depth of puncture using a simple mathematical principle. It also reviews the common problems faced during the process of puncture and dilatation and describes the ways to overcome them. The aim of this article is to provide the reader a step by step guide for percutaneous renal access. PMID:25789297

  12. Training femoral neck screw insertion skills to surgical trainees: computer-assisted surgery versus conventional fluoroscopic technique.

    PubMed

    Nousiainen, Markku T; Omoto, Daniel M; Zingg, Patrick O; Weil, Yoram A; Mardam-Bey, Sami W; Eward, William C

    2013-02-01

    : Femoral neck fractures are among the most common orthopaedic injuries impacting the health care system. Surgical management of such fractures with cannulated screws is a commonly performed procedure. The acquisition of surgical skills necessary to perform this procedure typically involves learning on real patients with fluoroscopic guidance. This study attempts to determine if a novel computer-navigated training model improves the learning of this basic surgical skill. A multicenter, prospective, randomized, and controlled study was conducted using surgical trainees with no prior experience in surgically managing femoral neck fractures. After a training session, participants underwent a pretest by performing the surgical task (screw placement) on a simulated hip fracture using fluoroscopic guidance. Immediately after, participants were randomized into either undergoing a training session using conventional fluoroscopy or computer-based navigation. Immediate posttest, retention (4 weeks later), and transfer tests were performed. Performance during the tests was determined by radiographic analysis of hardware placement. Screw placement by trainees was ultimately equal to the level of an expert surgeon with either training technique. Participants who trained with computer navigation took fewer attempts to position hardware and used less fluoroscopy time than those trained with fluoroscopy. When those trained with fluoroscopy used computer navigation at the transfer test, less fluoroscopy time and dosage was used. The concurrent augmented feedback provided by computer navigation did not affect the learning of this basic surgical skill in surgical novices. No compromise in learning occurred if the surgical novice trained with one type of technology and transferred to using the other. The findings of this study suggest that computer navigation may be safely used to train surgical novices in a basic procedure. This model avoids using both live patients and harmful radiation without a compromise in the acquisition of a 3-dimensional technical skill.

  13. CT Fluoroscopy-Guided Transsacral Intervertebral Drainage for Pyogenic Spondylodiscitis at the Lumbosacral Junction

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Matsumoto, Tomohiro, E-mail: t-matsu@tokai-u.jp; Mine, Takahiko, E-mail: mine@tsc.u-tokai.ac.jp; Hayashi, Toshihiko, E-mail: t.hayashi@tokai.ac.jp

    PurposeTo retrospectively describe the feasibility and efficacy of CT fluoroscopy-guided transsacral intervertebral drainage for pyogenic spondylodiscitis at the lumbosacral junction with a combination of two interventional radiological techniques—CT-guided bone biopsy and abscess drainage.Materials and methodsThree patients with pyogenic spondylodiscitis at the lumbosacral junction were enrolled in this study between July 2013 and December 2015. The procedure of CT fluoroscopy-guided transsacral intervertebral drainage for pyogenic spondylodiscitis at the lumbosacral junction was as follows: the sacrum at S1 pedicle was penetrated with an 11-gauge (G) bone biopsy needle to create a path for an 8-French (F) pigtail drainage catheter. The bone biopsymore » needle was withdrawn, and an 18-G needle was inserted into the intervertebral space of the lumbosacral junction. Then, a 0.038-inch guidewire was inserted into the intervertebral space. Finally, the 8-F pigtail drainage catheter was inserted over the guidewire until its tip reached the intervertebral space. All patients received six-week antibiotics treatment.ResultsSuccessful placement of the drainage catheter was achieved for each patient without procedural complications. The duration of drainage was 17–33 days. For two patients, specific organisms were isolated; thus, definitive medical therapy was possible. All patients responded well to the treatment.ConclusionsCT fluoroscopy-guided transsacral intervertebral drainage for pyogenic spondylodiscitis at the lumbosacral junction is feasible and can be effective with a combination of two interventional techniques—CT fluoroscopy-guided bone biopsy and abscess drainage.« less

  14. Iatrogenic deep musculocutaneous radiation injury following percutaneous coronary intervention.

    PubMed

    Monaco, JoAn L; Bowen, Kanika; Tadros, Peter N; Witt, Peter D

    2003-08-01

    Radiation-induced skin injury has been reported for multiple fluoroscopic procedures. Previous studies have indicated that prolonged fluoroscopic exposure during even a single percutaneous coronary intervention (PCI) may lead to cutaneous radiation injury. We document a novel case of deep muscle damage requiring wide local debridement and muscle flap reconstruction in a 59-year-old man with a large radiation-induced wound to the lower thoracic region following 1 prolonged PCI procedure. The deep muscular iatrogenic injury described in this report may be the source of significant morbidity. Recommendations to reduce radiation-induced damage include careful examination of the skin site before each procedure, minimized fluoroscopy time, utilization of pulse fluoroscopy, employment of radiation filters, and collimator s and rotation of the location of the image intensifier.

  15. Three-dimensional Image Fusion Guidance for Transjugular Intrahepatic Portosystemic Shunt Placement.

    PubMed

    Tacher, Vania; Petit, Arthur; Derbel, Haytham; Novelli, Luigi; Vitellius, Manuel; Ridouani, Fourat; Luciani, Alain; Rahmouni, Alain; Duvoux, Christophe; Salloum, Chady; Chiaradia, Mélanie; Kobeiter, Hicham

    2017-11-01

    To assess the safety, feasibility and effectiveness of image fusion guidance with pre-procedural portal phase computed tomography with intraprocedural fluoroscopy for transjugular intrahepatic portosystemic shunt (TIPS) placement. All consecutive cirrhotic patients presenting at our interventional unit for TIPS creation from January 2015 to January 2016 were prospectively enrolled. Procedures were performed under general anesthesia in an interventional suite equipped with flat panel detector, cone-beam computed tomography (CBCT) and image fusion technique. All TIPSs were placed under image fusion guidance. After hepatic vein catheterization, an unenhanced CBCT acquisition was performed and co-registered with the pre-procedural portal phase CT images. A virtual path between hepatic vein and portal branch was made using the virtual needle path trajectory software. Subsequently, the 3D virtual path was overlaid on 2D fluoroscopy for guidance during portal branch cannulation. Safety, feasibility, effectiveness and per-procedural data were evaluated. Sixteen patients (12 males; median age 56 years) were included. Procedures were technically feasible in 15 of the 16 patients (94%). One procedure was aborted due to hepatic vein catheterization failure related to severe liver distortion. No periprocedural complications occurred within 48 h of the procedure. The median dose-area product was 91 Gy cm 2 , fluoroscopy time 15 min, procedure time 40 min and contrast media consumption 65 mL. Clinical benefit of the TIPS placement was observed in nine patients (56%). This study suggests that 3D image fusion guidance for TIPS is feasible, safe and effective. By identifying virtual needle path, CBCT enables real-time multiplanar guidance and may facilitate TIPS placement.

  16. Feasibility and safety of transfemoral intra-arterial chemotherapy for head and neck cancer using a 3-French catheter system: comparison with a 4-French catheter system.

    PubMed

    Watanabe, Shigeru; Yamamoto, Akira; Torigoe, Teruyuki; Kanki, Akihiko; Tamada, Tsutomu; Ito, Katsuyoshi

    2016-02-01

    To assess the technical feasibility of transfemoral intra-arterial chemotherapy for head and neck cancer using a 3-French catheter system (3-Fr). Sixty-two patients with head and neck cancer who underwent transfemoral intra-arterial chemotherapy were included in this study. Thirty-three patients underwent treatment using a 3-Fr (group 3-Fr). Twenty-nine patients underwent treatment using a 4-French catheter system (group 4-Fr). The technical success rate, duration of the procedure with fluoroscopy, and rate of procedure-related complications were compared between group 3-Fr and group 4-Fr. In addition, in group 3-Fr, bleeding at the puncture site after 1.5 h of bed rest was evaluated. The technical success rate was 100% in both groups. The duration of the procedure with fluoroscopy didn't differ between group 3-Fr (mean 28.0 min) and group 4-Fr (mean 30.2 min) (p = 0.524). There was no procedure-related complication in either group. In group 3-Fr, no hemorrhagic complication was observed. A 3-French catheter system can be used to perform transfemoral intra-arterial chemotherapy for head and neck cancer and is technically feasible with approximately the same duration of the procedure with fluoroscopy. Furthermore, this method may shorten the bed rest time without hemorrhagic complication, and may reduce the risk of pulmonary embolism.

  17. Eye lens exposure to medical staff during endoscopic retrograde cholangiopancreatography.

    PubMed

    Zagorska, A; Romanova, K; Hristova-Popova, J; Vassileva, J; Katzarov, K

    2015-11-01

    The paper presents a study of the radiation doses to eye lens of medical staff during endoscopic retrograde cholangiopancreatography (ERCP) procedures performed in a busy gastroenterology department. For each procedure the dose equivalent to the eye, exposure time, dose rate, Kerma Area Product and fluoroscopy time were recorded. Measurements were performed for a period of two months in four main positions of the operating staff, and then extrapolated to estimate annual doses. The fluoroscopy time per ERCP procedure varied between 1.0 min and 28.8 min, with a mean value of 4.6 min. The calculated mean eye dose per procedure varied between 34.9 μSv and 93.3 μSv. The results demonstrated that if eye protection is not used, annual doses to the eye lens of the gastroenterologist performing the procedure and the anesthesiologist can exceed the dose limit of 20 mSv per year. Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  18. Comprehensive assessment of patient image quality and radiation dose in latest generation cardiac x-ray equipment for percutaneous coronary interventions

    PubMed Central

    Gislason-Lee, Amber J.; Keeble, Claire; Egleston, Daniel; Bexon, Josephine; Kengyelics, Stephen M.; Davies, Andrew G.

    2017-01-01

    Abstract. This study aimed to determine whether a reduction in radiation dose was found for percutaneous coronary interventional (PCI) patients using a cardiac interventional x-ray system with state-of-the-art image enhancement and x-ray optimization, compared to the current generation x-ray system, and to determine the corresponding impact on clinical image quality. Patient procedure dose area product (DAP) and fluoroscopy duration of 131 PCI patient cases from each x-ray system were compared using a Wilcoxon test on median values. Significant reductions in patient dose (p≪0.001) were found for the new system with no significant change in fluoroscopy duration (p=0.2); procedure DAP reduced by 64%, fluoroscopy DAP by 51%, and “cine” acquisition DAP by 76%. The image quality of 15 patient angiograms from each x-ray system (30 total) was scored by 75 clinical professionals on a continuous scale for the ability to determine the presence and severity of stenotic lesions; image quality scores were analyzed using a two-sample t-test. Image quality was reduced by 9% (p≪0.01) for the new x-ray system. This demonstrates a substantial reduction in patient dose, from acquisition more than fluoroscopy imaging, with slightly reduced image quality, for the new x-ray system compared to the current generation system. PMID:28491907

  19. Virtual Simulation in Enhancing Procedural Training for Fluoroscopy-guided Lumbar Puncture: A Pilot Study.

    PubMed

    Ali, Saad; Qandeel, Monther; Ramakrishna, Rishi; Yang, Carina W

    2018-02-01

    Fluoroscopy-guided lumbar puncture (FGLP) is a basic procedural component of radiology residency and neuroradiology fellowship training. Performance of the procedure with limited experience is associated with increased patient discomfort as well as increased radiation dose, puncture attempts, and complication rate. Simulation in health care is a developing field that has potential for enhancing procedural training. We demonstrate the design and utility of a virtual reality simulator for performing FGLP. An FGLP module was developed on an ImmersiveTouch platform, which digitally reproduces the procedural environment with a hologram-like projection. From computed tomography datasets of healthy adult spines, we constructed a 3-D model of the lumbar spine and overlying soft tissues. We assigned different physical characteristics to each tissue type, which the user can experience through haptic feedback while advancing a virtual spinal needle. Virtual fluoroscopy as well as 3-D images can be obtained for procedural planning and guidance. The number of puncture attempts, the distance to the target, the number of fluoroscopic shots, and the approximate radiation dose can be calculated. Preliminary data from users who participated in the simulation were obtained in a postsimulation survey. All users found the simulation to be a realistic replication of the anatomy and procedure and would recommend to a colleague. On a scale of 1-5 (lowest to highest) rating the virtual simulator training overall, the mean score was 4.3 (range 3-5). We describe the design of a virtual reality simulator for performing FGLP and present the initial experience with this new technique. Copyright © 2018 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

  20. Reducing intraoperative duration and ionising radiation exposure during the insertion of distal locking screws of intramedullary nails: a small-scale study comparing the current fluoroscopic method against radiation-free, electromagnetic navigation.

    PubMed

    Grimwood, Darren; Harvey-Lloyd, Jane

    2016-12-01

    Intramedullary nailing is the standard surgical treatment for mid-diaphyseal fractures of long bones; however, it is also a high radiation dose procedure. Distal locking is regularly cited as a demanding element of the procedure, and there remains a reliance on X-ray fluoroscopy to locate the distal holes. A recently developed electromagnetic navigation (EMN) system allows radiation-free distal locking, with a virtual on-screen image. To compare operative duration, fluoroscopy time and radiation dose when using EMN over fluoroscopy, for the distal locking of intramedullary nails. Consecutive patients with mid-diaphyseal fractures of the tibia and femur, treatable with intramedullary nails, were prospectively enrolled during a 9-month period. The sample consisted of 29 individuals, 19 under fluoroscopic guidance and 10 utilising EMN. Participants were allocated depending on the type of intramedullary nail used and surgeon's preference. These were further divided into tibial and femoral subcategories, relative to the fracture site. EMN reduced fluoroscopy time by 49 (p = 0.038) and 28 s during tibial and femoral nailings, respectively. Radiation dose was reduced by 18 cGy/cm 2 (p = 0.046) during tibial and 181 cGy/cm 2 during femoral nailings when utilising EMN. Operative duration was 11 min slower during tibial nailings using EMN, but 38 min faster in respect of femoral nailings. This study has evidenced statistically significant reductions in both fluoroscopy time and radiation dose when using EMN for the distal locking of intramedullary nails. It is expected that overall operative duration would also decrease in line with similar studies, with increased usage and a larger sample.

  1. Radiation Exposure of Interventional Radiologists During Computed Tomography Fluoroscopy-Guided Renal Cryoablation and Lung Radiofrequency Ablation: Direct Measurement in a Clinical Setting.

    PubMed

    Matsui, Yusuke; Hiraki, Takao; Gobara, Hideo; Iguchi, Toshihiro; Fujiwara, Hiroyasu; Kawabata, Takahiro; Yamauchi, Takatsugu; Yamaguchi, Takuya; Kanazawa, Susumu

    2016-06-01

    Computed tomography (CT) fluoroscopy-guided renal cryoablation and lung radiofrequency ablation (RFA) have received increasing attention as promising cancer therapies. Although radiation exposure of interventional radiologists during these procedures is an important concern, data on operator exposure are lacking. Radiation dose to interventional radiologists during CT fluoroscopy-guided renal cryoablation (n = 20) and lung RFA (n = 20) was measured prospectively in a clinical setting. Effective dose to the operator was calculated from the 1-cm dose equivalent measured on the neck outside the lead apron, and on the left chest inside the lead apron, using electronic dosimeters. Equivalent dose to the operator's finger skin was measured using thermoluminescent dosimeter rings. The mean (median) effective dose to the operator per procedure was 6.05 (4.52) μSv during renal cryoablation and 0.74 (0.55) μSv during lung RFA. The mean (median) equivalent dose to the operator's finger skin per procedure was 2.1 (2.1) mSv during renal cryoablation, and 0.3 (0.3) mSv during lung RFA. Radiation dose to interventional radiologists during renal cryoablation and lung RFA were at an acceptable level, and in line with recommended dose limits for occupational radiation exposure.

  2. Significant Radiation Dose Reduction in the Hybrid Operating Room Using a Novel X-ray Imaging Technology.

    PubMed

    van den Haak, R F F; Hamans, B C; Zuurmond, K; Verhoeven, B A N; Koning, O H J

    2015-10-01

    To prospectively quantify radiation dose change in aortoiliac endovascular procedures in the hybrid operating room (OR) for patients and medical staff with a novel X-ray imaging technology (ClarityIQ technology), and to assess whether procedure or fluoroscopy time or dose of iodinated contrast was affected. A prospective study including 138 patients was performed to compare radiation dose before and after installation of a novel X-ray imaging technology. Endovascular aneurysm repair (EVAR) was performed in 37 patients and an endovascular procedure for aortoiliac occlusive disease (AIOD) in 101. Patient radiation dose in air kerma (AK) and dose area product (DAP), patient demographics, and procedural data were recorded. Staff radiation dose was measured with real time personal dosimetry measurements. In both the EVAR and AIOD groups the reference system, ALX (AlluraXper FD20; Philips Healthcare, Best, the Netherlands), was compared with the upgraded X-ray system, CIQ (AlluraClarity FD20; Philips Healthcare). Procedure time, fluoroscopy time, and iodinated contrast dose were recorded. Patient radiation dose reduction in the EVAR group, in median AK, was 56% (ALX = 1,262.5 mGy; CIQ = 556.0 mGy [p < .01]); and in median DAP it was 57% (ALX = 224.4 Gycm(2) and CIQ = 95.8 Gycm(2) [p < .01]). Patient radiation dose reduction in the AIOD group, in median AK, was 76% (ALX = 1,011.0 mGy; CIQ = 248.0 mGy [p < .01]); and in median DAP it was 73% (ALX = 138.1 Gycm(2); CIQ = 38.0 Gycm(2) [p < .01]). Staff dose reduction in the EVAR group was 16% (ALX = 70.1 μSv; CIQ = 59.2 μSv [p = .43]) and in the AIOD group it was 69% (ALX = 96.2 μSv; CIQ = 30.1 μSv [p < .01]). There was no statistically significant difference between patient demographics, procedure time, fluoroscopy time, and iodinated contrast medium use in the two treatment groups before and after installation. A novel X-ray imaging technology in the hybrid OR suite resulted in a significant reduction of patient and staff radiation dose without affecting procedure length, fluoroscopy time, or use of contrast. Copyright © 2015 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

  3. Virtual bronchoscopic navigation without X-ray fluoroscopy to diagnose peripheral pulmonary lesions: a randomized trial.

    PubMed

    Asano, Fumihiro; Ishida, Takashi; Shinagawa, Naofumi; Sukoh, Noriaki; Anzai, Masaki; Kanazawa, Kenya; Tsuzuku, Akifumi; Morita, Satoshi

    2017-12-11

    Transbronchial biopsy for peripheral pulmonary lesions is generally performed under X-ray fluoroscopy. Virtual bronchoscopic navigation (VBN) is a method in which virtual images of the bronchial route to the lesion are produced based on CT images obtained before VBN, and the bronchoscope is guided using these virtual images, improving the diagnostic yield of peripheral pulmonary lesions. VBN has the possibility of eliminating the need for X-ray fluoroscopy in the bronchoscopic diagnosis of peripheral lesions. To determine whether VBN can be a substitute for X-ray fluoroscopy, a randomized multicenter trial (non-inferiority trial) was performed in VBN and X-ray fluoroscopy (XRF) -assisted groups. The non-inferiority margin in the VBN-assisted group compared with the XRF-assisted group was set at 15%. The subjects consisted of 140 patients with peripheral pulmonary lesions with a mean diameter > 3 cm. In the VBN-assisted group, the bronchoscope was guided to the lesion using a VBN system without X-ray fluoroscopy. In the XRF-assisted group, the same bronchoscope was guided to the lesion under X-ray fluoroscopy. Subsequently, in both groups, the lesion was visualized using endobronchial ultrasonography with a guide sheath (EBUS/GS), and biopsy was performed. In this serial procedure, X-ray fluoroscopy was not used in the VBNA group. The subjects of analysis consisted of 129 patients. The diagnostic yield was 76.9% (50/65) in the VBN-assisted group and 85.9% (55/64) in the XRF-assisted group. The difference in the diagnostic yield between the two groups was -9.0% (95% confidence interval: -22.3% ~ 4.3%). The non-inferiority of the VBN-assisted group could not be confirmed. The rate of visualizing lesions by EBUS was 95.4% (62/65) in the VBN-assisted group and 96.9% (62/64) in the XRF-assisted group, being high in both groups. On EBUS/GS, a bronchoscope and biopsy instruments may be guided to the lesions using VBN without X-ray fluoroscopy, but X-ray fluoroscopy is necessary to improve the accuracy of sample collection from lesions. During transbronchial biopsy for peripheral pulmonary lesions, VBN cannot be a substitute for X-ray fluoroscopy. UMIN-CTR (UMIN000001710); registered 16 February 2009.

  4. Evaluation of a new very low dose imaging protocol: feasibility and impact on X-ray dose levels in electrophysiology procedures

    PubMed Central

    Bourier, Felix; Reents, Tilko; Ammar-Busch, Sonia; Buiatti, Alessandra; Kottmaier, Marc; Semmler, Verena; Telishevska, Marta; Brkic, Amir; Grebmer, Christian; Lennerz, Carsten; Kolb, Christof; Hessling, Gabriele; Deisenhofer, Isabel

    2016-01-01

    Aims This study presents and evaluates the impact of a new lowest-dose fluoroscopy protocol (Siemens AG), especially designed for electrophysiology (EP) procedures, on X-ray dose levels. Methods and results From October 2014 to March 2015, 140 patients underwent an EP study on an Artis zee angiography system. The standard low-dose protocol was operated at 23 nGy (fluoroscopy) and at 120 nGy (cine-loop), the new lowest-dose protocol was operated at 8 nGy (fluoroscopy) and at 36 nGy (cine-loop). Procedural data, X-ray times, and doses were analysed in 100 complex left atrial and in 40 standard EP procedures. The resulting dose–area products were 877.9 ± 624.7 µGym² (n = 50 complex procedures, standard low dose), 199 ± 159.6 µGym² (n = 50 complex procedures, lowest dose), 387.7 ± 36.0 µGym² (n = 20 standard procedures, standard low dose), and 90.7 ± 62.3 µGym² (n = 20 standard procedures, lowest dose), P < 0.01. In the low-dose and lowest-dose groups, procedure times were 132.6 ± 35.7 vs. 126.7 ± 34.7 min (P = 0.40, complex procedures) and 72.3 ± 20.9 vs. 85.2 ± 44.1 min (P = 0.24, standard procedures), radiofrequency (RF) times were 53.8 ± 26.1 vs. 50.4 ± 29.4 min (P = 0.54, complex procedures) and 10.1 ± 9.9 vs. 12.2 ± 14.7 min (P = 0.60, standard procedures). One complication occurred in the standard low-dose and lowest-dose groups (P = 1.0). Conclusion The new lowest-dose imaging protocol reduces X-ray dose levels by 77% compared with the currently available standard low-dose protocol. From an operator standpoint, lowest X-ray dose levels create a different, reduced image quality. The new image quality did not significantly affect procedure or RF times and did not result in higher complication rates. Regarding radiological protection, operating at lowest-dose settings should become standard in EP procedures. PMID:26589627

  5. Real-time three dimensional CT and MRI to guide interventions for congenital heart disease and acquired pulmonary vein stenosis.

    PubMed

    Suntharos, Patcharapong; Setser, Randolph M; Bradley-Skelton, Sharon; Prieto, Lourdes R

    2017-10-01

    To validate the feasibility and spatial accuracy of pre-procedural 3D images to 3D rotational fluoroscopy registration to guide interventional procedures in patients with congenital heart disease and acquired pulmonary vein stenosis. Cardiac interventions in patients with congenital and structural heart disease require complex catheter manipulation. Current technology allows registration of the anatomy obtained from 3D CT and/or MRI to be overlaid onto fluoroscopy. Thirty patients scheduled for interventional procedures from 12/2012 to 8/2015 were prospectively recruited. A C-arm CT using a biplane C-arm system (Artis zee, VC14H, Siemens Healthcare) was acquired to enable 3D3D registration with pre-procedural images. Following successful image fusion, the anatomic landmarks marked in pre-procedural images were overlaid on live fluoroscopy. The accuracy of image registration was determined by measuring the distance between overlay markers and a reference point in the image. The clinical utility of the registration was evaluated as either "High", "Medium" or "None". Seventeen patients with congenital heart disease and 13 with acquired pulmonary vein stenosis were enrolled. Accuracy and benefit of registration were not evaluated in two patients due to suboptimal images. The distance between the marker and the actual anatomical location was 0-2 mm in 18 (64%), 2-4 mm in 3 (11%) and >4 mm in 7 (25%) patients. 3D3D registration was highly beneficial in 18 (64%), intermediate in 3 (11%), and not beneficial in 7 (25%) patients. 3D3D registration can facilitate complex congenital and structural interventions. It may reduce procedure time, radiation and contrast dose.

  6. Interventional spinal procedures guided and controlled by a 3D rotational angiographic unit.

    PubMed

    Pedicelli, Alessandro; Verdolotti, Tommaso; Pompucci, Angelo; Desiderio, Flora; D'Argento, Francesco; Colosimo, Cesare; Bonomo, Lorenzo

    2011-12-01

    The aim of this paper is to demonstrate the usefulness of 2D multiplanar reformatting images (MPR) obtained from rotational acquisitions with cone-beam computed tomography technology during percutaneous extra-vascular spinal procedures performed in the angiography suite. We used a 3D rotational angiographic unit with a flat panel detector. MPR images were obtained from a rotational acquisition of 8 s (240 images at 30 fps), tube rotation of 180° and after post-processing of 5 s by a local work-station. Multislice CT (MSCT) is the best guidance system for spinal approaches permitting direct tomographic visualization of each spinal structure. Many operators, however, are trained with fluoroscopy, it is less expensive, allows real-time guidance, and in many centers the angiography suite is more frequently available for percutaneous procedures. We present our 6-year experience in fluoroscopy-guided spinal procedures, which were performed under different conditions using MPR images. We illustrate cases of vertebroplasty, epidural injections, selective foraminal nerve root block, facet block, percutaneous treatment of disc herniation and spine biopsy, all performed with the help of MPR images for guidance and control in the event of difficult or anatomically complex access. The integrated use of "CT-like" MPR images allows the execution of spinal procedures under fluoroscopy guidance alone in all cases of dorso-lumbar access, with evident limitation of risks and complications, and without need for recourse to MSCT guidance, thus eliminating CT-room time (often bearing high diagnostic charges), and avoiding organizational problems for procedures that need, for example, combined use of a C-arm in the CT room.

  7. Yttrium-90 Resin Microsphere Radioembolization Using an Antireflux Catheter: An Alternative to Traditional Coil Embolization for Nontarget Protection

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Morshedi, Maud M., E-mail: maud.morshedi@my.rfums.org; Bauman, Michael, E-mail: mbauman@ucsd.edu; Rose, Steven C., E-mail: scrose@ucsd.edu

    2015-04-15

    PurposeSerious complications can result from nontarget embolization during yttrium-90 (Y-90) transarterial radioembolization. Hepatoenteric artery coil embolization has been traditionally performed to prevent nontarget radioembolization. The U.S. Food and Drug Administration–approved Surefire Infusion System (SIS) catheter, designed to prevent reflux, is an alternative to coils. The hypothesis that quantifiable SIS procedural parameters are comparable to coil embolization was tested.MethodsFourteen patients aged 36–79 years with colorectal, neuroendocrine, hepatocellular, and other predominantly bilobar hepatic tumors who underwent resin microsphere Y-90 radioembolization using only the SIS catheter (n = 7) versus only detachable coils (n = 7) for nontarget protection were reviewed retrospectively. Procedure time, fluoroscopy time, contrast dose,more » radiation dose, and cost were evaluated.ResultsMultivariate analysis identified significant cohort differences in the procedural parameters evaluated (F(10, 3) = 10.39, p = 0.04). Between-group comparisons of the pretreatment planning procedure in the SIS catheter group compared to the coil embolization group demonstrated a significant reduction in procedure time (102.6 vs. 192.1 min, respectively, p = 0.0004), fluoroscopy time (14.3 vs. 49.7 min, respectively, p = 0.0016), and contrast material dose (mean dose of 174.3 vs. 265.0 mL, respectively, p = 0.0098). Procedural parameters were not significantly different between the two groups during subsequent dose delivery procedures. Overall cost of combined first-time radioembolization procedures was significantly less in the SIS group ($4252) compared to retrievable coil embolization ($11,123; p = 0.001).ConclusionThe SIS catheter results in a reduction in procedure time, fluoroscopy time, and contrast material dose and may be an attractive cost-effective alternative to detachable coil embolization for prevention of nontarget radioembolization.« less

  8. Real-time fusion of coronary CT angiography with x-ray fluoroscopy during chronic total occlusion PCI.

    PubMed

    Ghoshhajra, Brian B; Takx, Richard A P; Stone, Luke L; Girard, Erin E; Brilakis, Emmanouil S; Lombardi, William L; Yeh, Robert W; Jaffer, Farouc A

    2017-06-01

    The purpose of this study was to demonstrate the feasibility of real-time fusion of coronary computed tomography angiography (CTA) centreline and arterial wall calcification with x-ray fluoroscopy during chronic total occlusion (CTO) percutaneous coronary intervention (PCI). Patients undergoing CTO PCI were prospectively enrolled. Pre-procedural CT scans were integrated with conventional coronary fluoroscopy using prototype software. We enrolled 24 patients who underwent CTO PCI using the prototype CT fusion software, and 24 consecutive CTO PCI patients without CT guidance served as a control group. Mean age was 66 ± 11 years, and 43/48 patients were men. Real-time CTA fusion during CTO PCI provided additional information regarding coronary arterial calcification and tortuosity that generated new insights into antegrade wiring, antegrade dissection/reentry, and retrograde wiring during CTO PCI. Overall CTO success rates and procedural outcomes remained similar between the two groups, despite a trend toward higher complexity in the fusion CTA group. This study demonstrates that real-time automated co-registration of coronary CTA centreline and calcification onto live fluoroscopic images is feasible and provides new insights into CTO PCI, and in particular, antegrade dissection reentry-based CTO PCI. • Real-time semi-automated fusion of CTA/fluoroscopy is feasible during CTO PCI. • CTA fusion data can be toggled on/off as desired during CTO PCI • Real-time CT calcium and centreline overlay could benefit antegrade dissection/reentry-based CTO PCI.

  9. Radiation exposure--do urologists take it seriously in Turkey?

    PubMed

    Söylemez, Haluk; Altunoluk, Bülent; Bozkurt, Yaşar; Sancaktutar, Ahmet Ali; Penbegül, Necmettin; Atar, Murat

    2012-04-01

    A questionnaire was administered to urologists to evaluate attitudes and behaviors about protection from radiation exposure during fluoroscopy guided endourological procedures. The questionnaire was e-mailed to 1,482 urologists, including urology residents, specialists and urologists holding all levels of academic degrees, between May and June 2011. The questionnaire administered to study participants was composed of demographic questions, and questions on radiation exposure frequency, and the use of dosimeters and flexible protective clothes. If a respondent reported not using dosimeters or protective clothes, additional questions asked for the reason. Of the 1,482 questionnaires 394 (26.58%) were returned, of which 363 had completed answers. A total of 307 physicians (84.58%) were exposed to ionizing radiation, of whom 79.61% stated that they perform percutaneous nephrolithotomy at the clinic. Fluoroscopy guidance was the initial choice of 96.19% of urologists during percutaneous nephrolithotomy. Despite the common use of lead aprons (75.24%) most urologists did not use dosimeters (73.94%), eyeglasses (76.95%) or gloves (66.67%) while 46.44% always used thyroid shields during fluoroscopy. When asked why they did not use protective clothing, the most common answers were that protective clothes are not ergonomic and not practical. Results clearly highlight the lack of use of ionizing radiation protection devices and dosimeters during commonly performed fluoroscopy guided endourological procedures among urologists in Turkey. Copyright © 2012 American Urological Association Education and Research, Inc. Published by Elsevier Inc. All rights reserved.

  10. Fluoroscopic radiation exposure: are we protecting ourselves adequately?

    PubMed

    Hoffler, C Edward; Ilyas, Asif M

    2015-05-06

    While traditional intraoperative fluoroscopy protection relies on thyroid shields and aprons, recent data suggest that the surgeon's eyes and hands receive more exposure than previously appreciated. Using a distal radial fracture surgery model, we examined (1) radiation exposure to the eyes, thyroid, chest, groin, and hands of a surgeon mannequin; (2) the degree to which shielding equipment can decrease exposure; and (3) how exposure varies with fluoroscopy unit size. An anthropomorphic model was fit with radiation-attenuating glasses, a thyroid shield, an apron, and gloves. "Exposed" thermoluminescent dosimeters overlaid the protective equipment at the eyes, thyroid, chest, groin, and index finger while "shielded" dosimeters were placed beneath the protective equipment. Fluoroscopy position and settings were standardized. The mini-c-arm milliampere-seconds were fixed based on the selection of the kilovolt peak (kVp). Three mini and three standard c-arms scanned a model of the patient's wrist continuously for fifteen minutes each. Ten dosimeter exposures were recorded for each c-arm. Hand exposure averaged 31 μSv/min (range, 22 to 48 μSv/min), which was 13.0 times higher than the other recorded exposures. Eye exposure averaged 4 μSv/min, 2.2 times higher than the mean thyroid, chest, and groin exposure. Gloves reduced hand exposure by 69.4%. Glasses decreased eye exposure by 65.6%. There was no significant difference in exposure between mini and standard fluoroscopy. Surgeons' hands receive the most radiation exposure during distal radial plate fixation under fluoroscopy. There was a small but insignificant difference in mean exposure between standard fluoroscopy and mini-fluoroscopy, but some standard units resulted in lower exposure than some mini-units. On the basis of these findings, we recommend routine protective equipment to mitigate exposure to surgeons' hands and eyes, in addition to the thyroid, chest, and groin, during fluoroscopy procedures. Copyright © 2015 by The Journal of Bone and Joint Surgery, Incorporated.

  11. Insertion of tunneled hemodialysis catheters without fluoroscopy.

    PubMed

    Motta Elias, Rosilene; da Silva Makida, Sonia Cristina; Abensur, Hugo; Martins Castro, Manuel Carlos; Affonso Moysés, Rosa Maria; Pereira, Benedito Jorge; Bueno de Oliveira, Rodrigo; Luders, Cláudio; Romão, João Egidio

    2010-01-01

    The tunneled cuffed catheter (TCC) is used as a bridge access for hemodialysis. Few prospective studies have been designed to evaluate conversion from non-tunneled to TCC without the use of fluoroscopy when performed by nephrologists. We performed an observational prospective cohort in incident patients receiving hemodialysis through a non-tunneled right jugular vein catheter. 130 procedures were performed in 122 patients (51+/-18 years). The success rate was 100%. There was a total of 26,546 catheter days. Ninety-one of the 130 catheters were removed during the study period. Life table analysis revealed primary patency rates of 92%, 82%, and 68% at 30, 60, and 120 days, respectively. Infection requiring catheter removal occurred at a frequency of 0.09 per 100 catheter days. Catheter malfunction requiring intervention occurred at a rate of 0.03 per 100 catheter days. Hypertension and duration of existing non-tunneled catheter of less than 2 weeks were independently associated with better TCC survival. The conversion from non-tunneled to TCC performed by nephrologists and without fluoroscopy may be safe by using the internal right jugular vein. The ideal time to do this procedure is within less than 2 weeks of existing non-tunneled catheter.

  12. Impact of a Novel Catheter Tracking System on Radiation Exposure during the Procedural Phases of Atrial Fibrillation and Flutter Ablation.

    PubMed

    Malliet, Nicolas; Andrade, Jason G; Khairy, Paul; Thanh, Hien Kiem Nguyen; Venier, Sandrine; Dubuc, Marc; Dyrda, Katia; Guerra, Peter; Mondésert, Blandine; Rivard, Léna; Tadros, Rafik; Talajic, Mario; Thibault, Bernard; Roy, Denis; Macle, Laurent

    2015-07-01

    Fluoroscopic guidance is used to position catheters during cardiac ablation. We evaluated the impact of a novel nonfluoroscopic sensor-guided electromagnetic navigation system (MG) on radiation exposure during catheter ablation of atrial fibrillation (AF) or atrial flutter (AFL). A total of 134 consecutive patients referred for ablation of AF (n = 44) or AFL (n = 90) ablation were prospectively enrolled. In one group the MG system was used for nonfluoroscopic catheter positioning, whereas in the conventional group standard fluoroscopy was utilized. Fluoroscopy times were assessed for each stage of procedure and total radiation exposure was quantified. Patient characteristics were similar between the groups. The procedural end point was achieved in all. Median (interquartile range [IQR]) fluoroscopy times were 12.5 minutes (7.6, 17.4) MG group versus 21.5 minutes (15.3, 23.0) conventional group (P < 0.0001) for AF ablation, and 0.8 minutes (0.4, 2.5) MG group versus 9.9 minutes (5.1, 22.5) conventional group (P < 0.0001) for AFL ablation. Median (IQR) total radiation exposure (μGy·m(2)) was 1,107 (906, 2,033) MG group versus 2,835 (1,688, 3,855) conventional group (P = 0.0001) for AF ablation, and 161 (65, 537) MG group versus 1,651 (796, 4,569) conventional group (P < 0.0001) for AFL ablation. No difference in total procedural time was seen. The use of a novel nonfluoroscopic catheter tracking system is associated with a significant reduction in radiation exposure during AF and AFL ablation (61% and 90% reduction, respectively). In the era of heightened awareness of the importance of radiation reduction, this system represents a safe and efficient tool to decrease radiation exposure during electrophysiological ablation procedures. ©2015 Wiley Periodicals, Inc.

  13. Electromagnetic tracking of flexible robotic catheters enables "assisted navigation" and brings automation to endovascular navigation in an in vitro study.

    PubMed

    Schwein, Adeline; Kramer, Benjamin; Chinnadurai, Ponraj; Virmani, Neha; Walker, Sean; O'Malley, Marcia; Lumsden, Alan B; Bismuth, Jean

    2018-04-01

    Combining three-dimensional (3D) catheter control with electromagnetic (EM) tracking-based navigation significantly reduced fluoroscopy time and improved robotic catheter movement quality in a previous in vitro pilot study. The aim of this study was to expound on previous results and to expand the value of EM tracking with a novel feature, assistednavigation, allowing automatic catheter orientation and semiautomatic vessel cannulation. Eighteen users navigated a robotic catheter in an aortic aneurysm phantom using an EM guidewire and a modified 9F robotic catheter with EM sensors at the tip of both leader and sheath. All users cannulated two targets, the left renal artery and posterior gate, using four visualization modes: (1) Standard fluoroscopy (control). (2) 2D biplane fluoroscopy showing real-time virtual catheter localization and orientation from EM tracking. (3) 2D biplane fluoroscopy with novel EM assisted navigation allowing the user to define the target vessel. The robotic catheter orients itself automatically toward the target; the user then only needs to advance the guidewire following this predefined optimized path to catheterize the vessel. Then, while advancing the catheter over the wire, the assisted navigation automatically modifies catheter bending and rotation in order to ensure smooth progression, avoiding loss of wire access. (4) Virtual 3D representation of the phantom showing real-time virtual catheter localization and orientation. Standard fluoroscopy was always available; cannulation and fluoroscopy times were noted for every mode and target cannulation. Quality of catheter movement was assessed by measuring the number of submovements of the catheter using the 3D coordinates of the EM sensors. A t-test was used to compare the standard fluoroscopy mode against EM tracking modes. EM tracking significantly reduced the mean fluoroscopy time (P < .001) and the number of submovements (P < .02) for both cannulation tasks. For the posterior gate, mean cannulation time was also significantly reduced when using EM tracking (P < .001). The use of novel EM assisted navigation feature (mode 3) showed further reduced cannulation time for the posterior gate (P = .002) and improved quality of catheter movement for the left renal artery cannulation (P = .021). These results confirmed the findings of a prior study that highlighted the value of combining 3D robotic catheter control and 3D navigation to improve safety and efficiency of endovascular procedures. The novel EM assisted navigation feature augments the robotic master/slave concept with automated catheter orientation toward the target and shows promising results in reducing procedure time and improving catheter motion quality. Copyright © 2017 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

  14. Does fluoroscopy improve outcomes in paediatric forearm fracture reduction?

    PubMed

    Menachem, S; Sharfman, Z T; Perets, I; Arami, A; Eyal, G; Drexler, M; Chechik, O

    2016-06-01

    To compare the radiographic results of paediatric forearm fracture reduced with and without fluoroscopic enhancement to investigate whether fractures reduced under fluoroscopic guidance would have smaller residual deformities and lower rates of re-reduction and surgery. A retrospective cohort analysis was conducted comparing paediatric patients with acute forearm fracture in two trauma centres. Demographics and radiographic data from paediatric forearm fractures treated in Trauma Centre A with the aid of a C-arm fluoroscopy were compared to those treated without fluoroscopy in Trauma Centre B. Re-reduction, late displacement, post-reduction deformity, and need for surgical intervention were compared between the two groups. The cohort included 229 children (175 boys and 54 girls, mean age 9.41±3.2 years, range 1-16 years) with unilateral forearm fractures (83 manipulated with fluoroscopy and 146 without). Thirty-four (15%) children underwent re-reduction procedures in the emergency department. Fifty-three (23%) children had secondary displacement in the cast, of which 18 were operated on, 20 were re-manipulated, and the remaining 15 were kept in the cast with an acceptable deformity. Twenty-nine additional children underwent operation for reasons other than secondary displacement. There were no significant differences in re-reduction and surgery rates or in post-reduction deformities between the two groups. The use of fluoroscopy during reduction of forearm fractures in the paediatric population apparently does not have a significant effect on patient outcomes. Reductions performed without fluoroscopy were comparably accurate in correcting deformities in both coronal and sagittal planes. Copyright © 2016 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.

  15. Radiation safety protocol using real-time dose reporting reduces patient exposure in pediatric electrophysiology procedures.

    PubMed

    Patel, Akash R; Ganley, Jamie; Zhu, Xiaowei; Rome, Jonathan J; Shah, Maully; Glatz, Andrew C

    2014-10-01

    Radiation exposure during pediatric catheterization is significant. We sought to describe radiation exposure and the effectiveness of radiation safety protocols in reducing exposure during catheter ablations with electrophysiology studies in children and patients with congenital heart disease. We additionally sought to identify at-risk patients. We retrospectively reviewed all interventional electrophysiology procedures performed from April 2009 to September 2011 (6 months preceding intervention, 12 months following implementation of initial radiation safety protocol, and 8 months following implementation of modified protocol). The protocols consisted of low pulse rate fluoroscopy settings, operator notification of skin entrance dose every 1,000 mGy, adjusting cameras by >5 at every 1,000 mGy, and appropriate collimation. The cohort consisted of 291 patients (70 pre-intervention, 137 after initial protocol implementation, 84 after modified protocol implementation) at a median age of 14.9 years with congenital heart disease present in 11 %. Diagnoses included atrioventricular nodal reentrant tachycardia (25 %), atrioventricular reentrant tachycardia (61 %), atrial tachycardias (12 %), and ventricular tachycardia (2 %). There were no differences between groups based on patient, arrhythmia, and procedural characteristics. Following implementation of the protocols, there were significant reductions in all measures of radiation exposure: fluoroscopy time (17.8 %), dose area product (80.2 %), skin entry dose (81.0 %), and effective dose (76.9 %), p = 0.0001. Independent predictors of increased radiation exposure included larger patient weight, longer fluoroscopy time, and lack of radiation safety protocol. Implementation of a radiation safety protocol for pediatric and congenital catheter ablations can drastically reduce radiation exposure to patients without affecting procedural success.

  16. Prospective Measurement of Patient Exposure to Radiation During Pediatric Ureteroscopy

    PubMed Central

    Kokorowski, Paul J.; Chow, Jeanne S.; Strauss, Keith; Pennison, Melanie; Routh, Jonathan C.; Nelson, Caleb P.

    2013-01-01

    Objective Little data have been reported regarding radiation exposure during pediatric endourologic procedures, including ureteroscopy (URS). We sought to measure radiation exposure during pediatric URS and identify opportunities for exposure reduction. Methods We prospectively observed URS procedures as part of a quality improvement initiative. Pre-operative patient characteristics, operative factors, fluoroscopy settings and radiation exposure were recorded. Our outcomes were entrance skin dose (ESD, in mGy) and midline dose (MLD, in mGy). Specific modifiable factors were identified as targets for potential quality improvement. Results Direct observation was performed on 56 consecutive URS procedures. Mean patient age was 14.8 ± 3.8 years (range 7.4 to 19.2); 9 children were under age 12 years. Mean ESD was 46.4 ± 48 mGy. Mean MLD was 6.2 ± 5.0 mGy. The most important major determinant of radiation dose was total fluoroscopy time (mean 2.68 ± 1.8 min) followed by dose rate setting, child anterior-posterior (AP) diameter, and source to skin distance (all p<0.01). The analysis of factors affecting exposure levels found that the use of ureteral access sheaths (p=0.01) and retrograde pyelography (p=0.04) were significantly associated with fluoroscopy time. We also found that dose rate settings were higher than recommended in up to 43% of cases and ideal C-arm positioning could have reduced exposure 14% (up to 49% in some cases). Conclusions Children receive biologically significant radiation doses during URS procedures. Several modifiable factors contribute to dose and could be targeted in efforts to implement dose reduction strategies. PMID:22341275

  17. Standardized methodology for transfemoral transcatheter aortic valve replacement with the Edwards Sapien XT valve under fluoroscopy guidance.

    PubMed

    Kasel, Albert M; Shivaraju, Anupama; Schneider, Stephan; Krapf, Stephan; Oertel, Frank; Burgdorf, Christof; Ott, Ilka; Sumer, Christian; Kastrati, Adnan; von Scheidt, Wolfgang; Thilo, Christian

    2014-09-01

    To provide a simplified, standardized methodology for a successful transfemoral transcatheter aortic valve replacement (TAVR) procedure with the Sapien XT valve in patients with severe aortic stenosis (AS). TAVR is currently reserved for patients with severe, symptomatic AS who are inoperable or at high operative risk. In many institutions, TAVR is performed under general anesthesia with intubation or with conscious sedation. In addition, many institutions still use transesophageal echo (TEE) during the procedure for aortic root angulations and positioning of the valve prior to implantation. Methods. We enrolled 100 consecutive patients (mean age, 80 ± 7 years; range, 50-94 years; female n=59) with severe symptomatic AS. Annulus measurements were based on computed tomography angiograms. All patients underwent fluoroscopy-guided transfemoral TAVR with little to no sedation and without simultaneous TEE. TAVR was predominantly performed with the use of local and central analgesics; only 36% of our cohort received conscious sedation. Procedural success of TAVR was 99%. Transthoracic echocardiography before discharge excluded aortic regurgitation (AR) >2 in all patients (AR >1; n=6). In-hospital stroke rate was 6%. The vessel closure system was successfully employed in 96%. Major vascular complication rate was 1%. The 30-day mortality was 2%. Fluoroscopy-guided TAVR with the use of just analgesics with or without conscious sedation is safe and effective, and this potentially enables a more time-effective and cost-effective procedure. This paper provides simplified, stepwise guidance on how to perform transfemoral TAVR with the Sapien XT valve.

  18. Low dose tomographic fluoroscopy: 4D intervention guidance with running prior

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Flach, Barbara; Kuntz, Jan; Brehm, Marcus

    Purpose: Today's standard imaging technique in interventional radiology is the single- or biplane x-ray fluoroscopy which delivers 2D projection images as a function of time (2D+T). This state-of-the-art technology, however, suffers from its projective nature and is limited by the superposition of the patient's anatomy. Temporally resolved tomographic volumes (3D+T) would significantly improve the visualization of complex structures. A continuous tomographic data acquisition, if carried out with today's technology, would yield an excessive patient dose. Recently the authors proposed a method that enables tomographic fluoroscopy at the same dose level as projective fluoroscopy which means that if scanning time ofmore » an intervention guided by projective fluoroscopy is the same as that of an intervention guided by tomographic fluoroscopy, almost the same dose is administered to the patient. The purpose of this work is to extend authors' previous work and allow for patient motion during the intervention.Methods: The authors propose the running prior technique for adaptation of a prior image. This adaptation is realized by a combination of registration and projection replacement. In a first step the prior is deformed to the current position via affine and deformable registration. Then the information from outdated projections is replaced by newly acquired projections using forward and backprojection steps. The thus adapted volume is the running prior. The proposed method is validated by simulated as well as measured data. To investigate motion during intervention a moving head phantom was simulated. Real in vivo data of a pig are acquired by a prototype CT system consisting of a flat detector and a continuously rotating clinical gantry.Results: With the running prior technique it is possible to correct for motion without additional dose. For an application in intervention guidance both steps of the running prior technique, registration and replacement, are necessary. Reconstructed volumes based on the running prior show high image quality without introducing new artifacts and the interventional materials are displayed at the correct position.Conclusions: The running prior improves the robustness of low dose 3D+T intervention guidance toward intended or unintended patient motion.« less

  19. Efficacy of RADPAD® protection drape in reducing radiation exposure to the primary operator during Transcatheter Aortic Valve Implantation (TAVI).

    PubMed

    Sharma, Divyesh; Ramsewak, Adesh; Manoharan, Ganesh; Spence, Mark S

    2016-02-01

    The efficacy of RADPAD® (a sterile, lead-free drape) has been demonstrated to reduce the scatter radiation to the primary operator during fluoroscopic procedures. However, the use of the RADPAD® during TAVI procedures has not been studied. Transcatheter aortic valve implantation (TAVI) is now an established treatment for patients with symptomatic severe aortic stenosis who are deemed inoperable or at high risk for conventional surgical aortic valve replacement (AVR). Consequently the radiation exposure to the patient and the interventional team from this procedure has become a matter of interest and importance. Methods to reduce radiation exposure to the interventional team during this procedure should be actively investigated. In this single center prospective study, we determined the radiation dose during this procedure and the efficacy of RADPAD® in reducing the radiation dose to the primary operator. Fifty consecutive patients due to undergo elective TAVI procedures were identified. Patients were randomly assigned to undergo the procedure with or without the use of a RADPAD® drape. There were 25 patients in each group and dosimetry was performed at the left eye level of the primary operator. The dosimeter was commenced at the start of the procedure, and the dose was recorded immediately after the end of the procedure. Fluoroscopy times and DAP were also recorded prospectively. Twenty-five patients underwent transfemoral TAVI using a RADPAD® and 25 with no-RADPAD®. The mean primary operator radiation dose was significantly lower in the RADPAD group at 14.8 mSv vs. 24.3 mSv in the no-RADPAD group (P=0.008). There was no significant difference in fluoroscopy times or dose-area products between the two patient groups. The dose to the primary operator relative to fluoroscopy time (RADPAD: slope=0.325; no RADPAD: slope=1.148; analysis of covariance F=7.47, P=0.009) and dose area product (RADPAD: slope=0.0007; no RADPAD: slope=0.002; analysis of covariance F=7.38; P=0.009) was smaller in the RADPAD group compared to no-RADPAD group. Use of a RADPAD® significantly reduces radiation exposure to the primary operator during TAVI procedures.

  20. Evaluation of a new very low dose imaging protocol: feasibility and impact on X-ray dose levels in electrophysiology procedures.

    PubMed

    Bourier, Felix; Reents, Tilko; Ammar-Busch, Sonia; Buiatti, Alessandra; Kottmaier, Marc; Semmler, Verena; Telishevska, Marta; Brkic, Amir; Grebmer, Christian; Lennerz, Carsten; Kolb, Christof; Hessling, Gabriele; Deisenhofer, Isabel

    2016-09-01

    This study presents and evaluates the impact of a new lowest-dose fluoroscopy protocol (Siemens AG), especially designed for electrophysiology (EP) procedures, on X-ray dose levels. From October 2014 to March 2015, 140 patients underwent an EP study on an Artis zee angiography system. The standard low-dose protocol was operated at 23 nGy (fluoroscopy) and at 120 nGy (cine-loop), the new lowest-dose protocol was operated at 8 nGy (fluoroscopy) and at 36 nGy (cine-loop). Procedural data, X-ray times, and doses were analysed in 100 complex left atrial and in 40 standard EP procedures. The resulting dose-area products were 877.9 ± 624.7 µGym² (n = 50 complex procedures, standard low dose), 199 ± 159.6 µGym² (n = 50 complex procedures, lowest dose), 387.7 ± 36.0 µGym² (n = 20 standard procedures, standard low dose), and 90.7 ± 62.3 µGym² (n = 20 standard procedures, lowest dose), P < 0.01. In the low-dose and lowest-dose groups, procedure times were 132.6 ± 35.7 vs. 126.7 ± 34.7 min (P = 0.40, complex procedures) and 72.3 ± 20.9 vs. 85.2 ± 44.1 min (P = 0.24, standard procedures), radiofrequency (RF) times were 53.8 ± 26.1 vs. 50.4 ± 29.4 min (P = 0.54, complex procedures) and 10.1 ± 9.9 vs. 12.2 ± 14.7 min (P = 0.60, standard procedures). One complication occurred in the standard low-dose and lowest-dose groups (P = 1.0). The new lowest-dose imaging protocol reduces X-ray dose levels by 77% compared with the currently available standard low-dose protocol. From an operator standpoint, lowest X-ray dose levels create a different, reduced image quality. The new image quality did not significantly affect procedure or RF times and did not result in higher complication rates. Regarding radiological protection, operating at lowest-dose settings should become standard in EP procedures. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2015. For permissions please email: journals.permissions@oup.com.

  1. Comparison of measured and estimated maximum skin doses during CT fluoroscopy lung biopsies.

    PubMed

    Zanca, F; Jacobs, A; Crijns, W; De Wever, W

    2014-07-01

    To measure patient-specific maximum skin dose (MSD) associated with CT fluoroscopy (CTF) lung biopsies and to compare measured MSD with the MSD estimated from phantom measurements, as well as with the CTDIvol of patient examinations. Data from 50 patients with lung lesions who underwent a CT fluoroscopy-guided biopsy were collected. The CT protocol consisted of a low-kilovoltage (80 kV) protocol used in combination with an algorithm for dose reduction to the radiology staff during the interventional procedure, HandCare (HC). MSD was assessed during each intervention using EBT2 gafchromic films positioned on patient skin. Lesion size, position, total fluoroscopy time, and patient-effective diameter were registered for each patient. Dose rates were also estimated at the surface of a normal-size anthropomorphic thorax phantom using a 10 cm pencil ionization chamber placed at every 30°, for a full rotation, with and without HC. Measured MSD was compared with MSD values estimated from the phantom measurements and with the cumulative CTDIvol of the procedure. The median measured MSD was 141 mGy (range 38-410 mGy) while the median cumulative CTDIvol was 72 mGy (range 24-262 mGy). The ratio between the MSD estimated from phantom measurements and the measured MSD was 0.87 (range 0.12-4.1) on average. In 72% of cases the estimated MSD underestimated the measured MSD, while in 28% of the cases it overestimated it. The same trend was observed for the ratio of cumulative CTDIvol and measured MSD. No trend was observed as a function of patient size. On average, estimated MSD from dose rate measurements on phantom as well as from CTDIvol of patient examinations underestimates the measured value of MSD. This can be attributed to deviations of the patient's body habitus from the standard phantom size and to patient positioning in the gantry during the procedure.

  2. Comparison Perioperative Factors During Minimally Invasive Pre-Psoas Lateral Interbody Fusion of the Lumbar Spine Using Either Navigation or Conventional Fluoroscopy

    PubMed Central

    Zhang, Yue-Hui; White, Ian; Potts, Eric; Mobasser, Jean-Pierre

    2017-01-01

    Study Design: Retrospective clinical study. Objectives: The aim of this study was to compare intraoperative conditions and clinical results of patients undergoing pre-psoas oblique lateral interbody fusion (OLIF) using navigation or conventional fluoroscopy (C-ARM) techniques. Methods: Forty-two patients (22 patients by navigation and 20 by fluoroscopy) underwent the OLIF procedure at 2 medical centers, and records were reviewed. Clinical data was collected and compared between the 2 groups. Patients were followed-up with a range of 6 to 24 months. Results: There were no significant differences on demographic data between groups. The navigation group had zero radiation exposure (RE) to the surgeon and radiation time compared to the C-ARM group, with total RE of 44.59 ± 26.65 mGy and radiation time of 88.30 ± 58.28 seconds (P < .05). The RE to the patient was significantly lower in the O-ARM group (9.38 mGy) compared to the C-ARM group (44.59 ± 26.65 mGy). Operating room time was slightly longer in the navigation group (2.49 ± 1.35 hours) compared to the C-ARM group (2.30 ± 1.17 hours; P > .05), although not statistically significant. No differences were found in estimated blood loss, length of hospitalization, surgery-related complications, and outcome scores with an average of 8-month follow-up. Conclusions: Compared with C-ARM techniques, using navigation can eliminate RE to surgeon and decrease RE to the patient, and it had no significant effect on operating time, estimated blood loss, length of hospitalization, or perioperative complications in the patients with OLIF procedure. This study shows that navigation is a safe alternative to fluoroscopy during the OLIF procedure in the treatment of degenerative lumbar conditions. PMID:28989845

  3. Multicenter, randomized comparison between magnetically navigated and manually guided radiofrequency ablation of atrioventricular nodal reentrant tachycardia (the MagMa-AVNRT-trial).

    PubMed

    Reents, Tilko; Jilek, Clemens; Schuster, Peter; Nölker, Georg; Koch-Büttner, Katharina; Ammar-Busch, Sonia; Semmler, Verena; Bourier, Felix; Kottmaier, Marc; Kornmayer, Marie; Brooks, Stephanie; Fichtner, Stephanie; Kolb, Christof; Deisenhofer, Isabel; Hessling, Gabriele

    2017-12-01

    Remote magnetic navigation (RMN) is attributed to diminish radiation exposure for both patient and operator performing catheter ablation for different arrhythmia substrates. The purpose of this prospective, randomized study was to compare RMN with manually guided catheter ablation for AV nodal reentrant tachycardia (AVNRT) regarding fluoroscopy time/dosage, acute and long-term efficacy as well as safety. A total of 218 patients with AVNRT undergoing catheter ablation at three centers (male 34%, mean age 50 ± 17 years) were randomized to a manual approach (n = 113) or RMN (n = 105) using the Niobe ® magnetic navigation system. The primary study endpoint was total fluoroscopy time/dosage for patient and operator at the end of the procedure. Secondary endpoints included acute success, procedure duration, complications and success rate after 6 months. Fluoroscopy time and dosage for the patient were significantly reduced in the RMN group compared to the manual group (6 ± 6 vs. 11 ± 10 min; p < 0.001 and 425 ± 558 vs. 751 ± 900 cGycm 2 , p = 0.002). A reduction in fluoroscopy time/dose also applied to the operator (3 ± 5 vs. 7 ± 9 min 209 ± 444 vs. 482 ± 689 cGycm 2 , p < 0.001). Procedure duration was significantly longer in the RMN group (88 ± 29 vs. 79 ± 29 min; p = 0.03) and crossover from the RMN group to manual ablation occurred in 7.6% of patients (7.6 vs. 0.1%; p = 0.02). Acute success was achieved in 100% of patients in both groups. Midterm success after 6 months was 97 vs. 98% (p = 0.67). No complications occurred in both groups. The use of RMN for catheter ablation of AVNRT compared to a manual approach results in a reduction of fluoroscopy time and dosage of about 50% for both patients and physicians. Acute and midterm success and safety are comparable. RMN is a good alternative to a manual approach for AVNRT ablation.

  4. Percutaneous Facet Screw Fixation in the Treatment of Symptomatic Recurrent Lumbar Facet Joint Cyst: A New Technique

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Amoretti, Nicolas, E-mail: amorettinicolas@yahoo.fr; Gallo, Giacomo, E-mail: giacomo.gallo83@gmail.com; Bertrand, Anne-Sophie, E-mail: asbertrand3@hotmail.com

    We present a case of percutaneous treatment of symptomatic recurrent lumbar facet joint cyst resistant to all medical treatments including facet joint steroid injection. Percutaneous transfacet fixation was then performed at L4–L5 level with a cannulated screw using CT and fluoroscopy guidance. The procedure time was 30 min. Using the visual analog scale (VAS), pain decreased from 9.5, preoperatively, to 0 after the procedure. At 6-month follow-up, an asymptomatic cystic recurrence was observed, which further reduced at the 1-year follow-up. Pain remained stable (VAS at 0) during all follow-ups. CT- and fluoroscopy-guided percutaneous cyst rupture associated with facet screw fixation couldmore » be an alternative to surgery in patients suffering from a symptomatic recurrent lumbar facet joint cyst.« less

  5. Retrieval characteristics of the Bard Denali and Argon Option inferior vena cava filters.

    PubMed

    Dowell, Joshua D; Semaan, Dominic; Makary, Mina S; Ryu, John; Khayat, Mamdouh; Pan, Xueliang

    2017-11-01

    The purpose of this study was to compare the retrieval characteristics of the Option Elite (Argon Medical, Plano, Tex) and Denali (Bard, Tempe, Ariz) retrievable inferior vena cava filters (IVCFs), two filters that share a similar conical design. A single-center, retrospective study reviewed all Option and Denali IVCF removals during a 36-month period. Attempted retrievals were classified as advanced if the routine "snare and sheath" technique was initially unsuccessful despite multiple attempts or an alternative endovascular maneuver or access site was used. Patient and filter characteristics were documented. In our study, 63 Option and 45 Denali IVCFs were retrieved, with an average dwell time of 128.73 and 99.3 days, respectively. Significantly higher median fluoroscopy times were experienced in retrieving the Option filter compared with the Denali filter (12.18 vs 6.85 minutes; P = .046). Use of adjunctive techniques was also higher in comparing the Option filter with the Denali filter (19.0% vs 8.7%; P = .079). No significant difference was noted between these groups in regard to gender, age, or history of malignant disease. Option IVCF retrieval procedures required significantly longer retrieval fluoroscopy time compared with Denali IVCFs. Although procedure time was not analyzed in this study, as a surrogate, the increased fluoroscopy time may also have an impact on procedural direct costs and throughput. Copyright © 2017 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

  6. Fluoroscopy Learning Curve in Hip Arthroscopy-A Single Surgeon's Experience.

    PubMed

    Smith, Kevin M; Duplantier, Neil L; Crump, Kimbelyn H; Delgado, Domenica A; Sullivan, Stephanie L; McCulloch, Patrick C; Harris, Joshua D

    2017-10-01

    To determine if (1) absorbed radiation dose and (2) fluoroscopy time decreased with experience over the first 100 cases of a single surgeon's hip arthroscopy practice. Subjects who underwent hip arthroscopy for symptomatic femoroacetabular impingement and labral injury were eligible for analysis. Inclusion criteria included the first 100 subjects who underwent hip arthroscopy by a single surgeon (December 2013 to December 2014). Subject demographics, procedure details, fluoroscopy absorbed dose (milligray [mGy]), and time were recorded. Subjects were categorized by date of surgery to one of 4 possible groups (25 per group). One-way analysis of variance was used to determine if a significant difference in dose (mGy) or time was present between groups. Simple linear regression analysis was performed to determine the relation between case number and both radiation dose and fluoroscopy time. Subjects underwent labral repair (n = 93), cam osteoplasty (n = 90), and pincer acetabuloplasty (n = 65). There was a significant (P < .001 for both) linear regression between case number and both radiation dose and fluoroscopy time. A significant difference in mGy was observed between groups, group 1 the highest and group 4 the lowest amounts of radiation (P = .003). Comparing individual groups, group 4 was found to have a significantly lower amount of radiation than group 1 (P = .002), though it was not significantly lower than that of group 2 (P = .09) or group 3 (P = .08). A significant difference in fluoroscopy time was observed between groups, group 1 the highest and group 4 the lowest times (P = .05). Comparing individual groups, group 4 was found to have a significantly lower fluoroscopy time than group 1 (P = .039). Correction for weight, height, and body mass index all revealed the same findings: significant (P < .05) differences in both dose and time across groups. The absorbed dose of radiation and fluoroscopy time decreased significantly over the first 100 cases of a single surgeon's hip arthroscopy practice learning curve. Level IV, therapeutic, retrospective, noncomparative case series. Copyright © 2017 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

  7. Risk-benefit analysis of navigation techniques for vertebral transpedicular instrumentation: a prospective study.

    PubMed

    Noriega, David C; Hernández-Ramajo, Rubén; Rodríguez-Monsalve Milano, Fiona; Sanchez-Lite, Israel; Toribio, Borja; Ardura, Francisco; Torres, Ricardo; Corredera, Raul; Kruger, Antonio

    2017-01-01

    Pedicle screws in spinal surgery have allowed greater biomechanical stability and higher fusion rates. However, malposition is very common and may cause neurologic, vascular, and visceral injuries and compromise mechanical stability. The purpose of this study was to compare the malposition rate between intraoperative computed tomography (CT) scan assisted-navigation and free-hand fluoroscopy-guided techniques for placement of pedicle screw instrumentation. This is a prospective, randomized, observational study. A total of 114 patients were included: 58 in the assisted surgery group and 56 in the free-hand fluoroscopy-guided surgery group. Analysis of screw position was assessed using the Heary classification. Breach severity was defined according to the Gertzbein classification. Radiation doses were evaluated using thermoluminescent dosimeters, and estimates of effective and organ doses were made based on scan technical parameters. Consecutive patients with degenerative disease, who underwent surgical procedures using the free-hand, or intraoperative navigation technique for placement of transpedicular instrumentation, were included in the study. Forty-four out of 625 implanted screws were malpositioned: 11 (3.6%) in the navigated surgery group and 33 (10.3%) in the free-hand group (p<.001). Screw position according to the Heary scale was Grade II (4 navigated surgery, 6 fluoroscopy guided), Grade III (3 navigated surgery, 11 fluoroscopy guided), Grade IV (4 navigated surgery, 16 fluoroscopy guided), and Grade V (1 fluoroscopy guided). There was only one symptomatic case in the conventional surgery group. Breach severity was seven Grade A and four Grade B in the navigated surgery group, and eight Grade A, 24 Grade B, and one Grade C in free-hand fluoroscopy-guided surgery group. Radiation received per patient was 5.8 mSv (4.8-7.3). The median dose received in the free-hand fluoroscopy group was 1 mGy (0.8-1.1). There was no detectable radiation level in the navigation-assisted surgery group, whereas the effective dose was 10 µGy in the free-hand fluoroscopy-guided surgery group. Malposition rate, both symptomatic and asymptomatic, in spinal surgery is reduced when using CT-guided placement of transpedicular instrumentation compared with placement under fluoroscopic guidance, with radiation values within the safety limits for health. Larger studies are needed to determine risk-benefit in these patients. Copyright © 2016 Elsevier Inc. All rights reserved.

  8. Editorial: ERCP-Related Radiation Cataractogenesis: Is It Time to Be Concerned?

    PubMed

    Mekaroonkamol, Parit; Keilin, Steven

    2017-05-01

    With the growing number of fluoroscopic guided endoscopic procedures, radiation-related risk needs to be further assessed. Recent evidence indicates that radiation cataractogenesis occurs at a lower dose threshold than previously believed. While body aprons and thyroid shields are well-established standard protection during fluoroscopy, ocular safety and the use of protective eyewear are not as well defined. This prospective study answered two important questions: Does the standard body dosimeter provide an accurate ocular dosimetry? And what is the time of fluoroscopy needed to warrant using lens protection? It also raises the question whether current guidelines need to be updated.

  9. Effectiveness of the implementation of a simple radiation reduction protocol in the catheterization laboratory.

    PubMed

    Jurado-Román, Alfonso; Sánchez-Pérez, Ignacio; Lozano Ruíz-Poveda, Fernando; López-Lluva, María T; Pinilla-Echeverri, Natalia; Moreno Arciniegas, Andrea; Agudo-Quilez, Pilar; Gil Agudo, Antonio

    2016-01-01

    A reduction in radiation doses at the catheterization laboratory, maintaining the quality of procedures is essential. Our objective was to analyze the results of a simple radiation reduction protocol at a high-volume interventional cardiology unit. We analyzed 1160 consecutive procedures: 580 performed before the implementation of the protocol and 580 after it. The protocol consisted in: the reduction of the number of ventriculographies and aortographies, the optimization of the collimation and the geometry of the X ray tube-patient-receptor, the use of low dose-rate fluoroscopy and the reduction of the number of cine sequences using the software "last fluoroscopy hold". There were no significant differences in clinical baseline features or in the procedural characteristics with the exception of a higher percentage of radial approach (30.7% vs 69.6%; p<0.001) and of percutaneous coronary interventions of chronic total occlusions after the implementation of the protocol (2.1% vs 6.7%; p=0,001). Angiographic success was similar during both periods (98.3% vs 99.2%; p=0.2). There were no significant differences between both periods regarding the overall duration of the procedures (26.9 vs 29.6min; p=0.14), or the fluoroscopy time (13.3 vs 13.2min; p=0.8). We observed a reduction in the percentage of procedures with ventriculography (80.9% vs 7.1%; p<0.0001) or aortography (15.4% vs 4.4%; p<0.0001), the cine runs (21.8 vs 6.9; p<0.0001) and the dose-area product (165 vs 71 Gyxcm(2); p<0.0001). With the implementation of a simple radiation reduction protocol, a 57% reduction of dose-area product was observed without a reduction in the quality or the complexity of procedures. Copyright © 2016 Elsevier Inc. All rights reserved.

  10. Optimizing radiofrequency ablation of paroxysmal and persistent atrial fibrillation by direct catheter force measurement-a case-matched comparison in 198 patients.

    PubMed

    Sigmund, Elisabeth; Puererfellner, Helmut; Derndorfer, Michael; Kollias, Georgios; Winter, Siegmund; Aichinger, Josef; Nesser, Hans-Joachim; Martinek, Martin

    2015-02-01

    Sufficient electrode-tissue contact is crucial for adequate lesion formation in radiofrequency catheter ablation (RFCA). We assessed the impact of direct catheter force measurement on acute procedural parameters and outcome of RFCA for paroxysmal and persistent atrial fibrillation (AF). Ninety-nine consecutive patients (70% men) with paroxysmal (63.6%) or persistent AF underwent left atrial RFCA using a 3.5-mm open-irrigated-tip (OIT) catheter with contact force measurement capabilities (group 1). For comparison a case-matched cohort with standard OIT catheters was used (99 patients; group 2). Case matching included gender, type of AF, number or RFCA procedures, and type of procedure. Procedural data showed a significant decline in radiofrequency ablation time from 52 ± 20 to 44 ± 16 minutes (P = 0.003) with a remarkable mean reduction in overall procedure time of 34 minutes (P = 0.0001; 225.8 ± 53.1 vs 191.9 ± 53.3 minutes). In parallel, the total fluoroscopy time could be significantly reduced from 28.5 ± 11.0 to 19.9 ± 9.3 minutes (P = 0.0001) as well as fluoroscopy dose from 74.1 ± 58.0 to 56.7 ± 38.9 Gy/cm(2) (P = 0.016). Periprocedural complications were similar in both groups. The use of contact force sensing technology is able to significantly reduce ablation, procedure, and fluoroscopy times as well as dose in RFCA of AF in a mixed case-matched group of paroxysmal and persistent AF. Energy delivery is substantially reduced by avoiding radiofrequency ablation in positions with insufficient surface contact. Additionally 12-month outcome data showed increased efficacy. Such time saving and equally safe technology may have a relevant impact on laboratory management and increased cost effectiveness. © 2014 Wiley Periodicals, Inc.

  11. Fiber sensor assisted in-vivo needle guidance for minimally invasive procedures (Conference Presentation)

    NASA Astrophysics Data System (ADS)

    Baghdadchi, Saharnaz; Chao, Cherng; Esener, Sadik; Mattrey, Robert F.; Eghtedari, Mohammad A.

    2017-02-01

    Image-guided procedures are performed frequently by radiologists to insert a catheter within a target vessel or lumen or to perform biopsy of a lesion. For instance, an interventional radiologist uses fluoroscopy during percutaneous biliary drainage procedure (a procedure during which a catheter is inserted through the skin to drain the bile from liver) to identify the location of the needle tip within liver parenchyma, hepatic blood vessel or bile duct. However, the identification of the target organ under fluoroscopy exposes the patient to x-ray irradiation, which may be significant if the time of procedure is prolonged. We have designed a fiber core needle system that may help the radiologist identify the location of the needle tip in real time without exposing the patient to x-ray. Our needle system transmits a low power modulated light into the tissue through a fiber cable embedded in the needle and detects the backscattered light using another fiber inside the needle. We were able to successfully distinguish the location of our prototype needle tip inside a cow liver phantom to identify if the needle tip was within liver parenchyma, liver vessels, or in the bile duct based on the recorded backscattered light.

  12. Echocardiographic and Fluoroscopic Fusion Imaging for Procedural Guidance: An Overview and Early Clinical Experience.

    PubMed

    Thaden, Jeremy J; Sanon, Saurabh; Geske, Jeffrey B; Eleid, Mackram F; Nijhof, Niels; Malouf, Joseph F; Rihal, Charanjit S; Bruce, Charles J

    2016-06-01

    There has been significant growth in the volume and complexity of percutaneous structural heart procedures in the past decade. Increasing procedural complexity and accompanying reliance on multimodality imaging have fueled the development of fusion imaging to facilitate procedural guidance. The first clinically available system capable of echocardiographic and fluoroscopic fusion for real-time guidance of structural heart procedures was approved by the US Food and Drug Administration in 2012. Echocardiographic-fluoroscopic fusion imaging combines the precise catheter and device visualization of fluoroscopy with the soft tissue anatomy and color flow Doppler information afforded by echocardiography in a single image. This allows the interventionalist to perform precise catheter manipulations under fluoroscopy guidance while visualizing critical tissue anatomy provided by echocardiography. However, there are few data available addressing this technology's strengths and limitations in routine clinical practice. The authors provide a critical review of currently available echocardiographic-fluoroscopic fusion imaging for guidance of structural heart interventions to highlight its strengths, limitations, and potential clinical applications and to guide further research into value of this emerging technology. Copyright © 2016 American Society of Echocardiography. Published by Elsevier Inc. All rights reserved.

  13. Intussusception reduction: Effect of air vs. liquid enema on radiation dose.

    PubMed

    Kaplan, Summer L; Magill, Dennise; Felice, Marc A; Edgar, J Christopher; Anupindi, Sudha A; Zhu, Xiaowei

    2017-10-01

    Both air and radiopaque liquid contrast are used to reduce ileocolic intussusception under fluoroscopy. Some suggest air lowers radiation dose due to shorter procedure times. However, air enema likely lowers radiation dose regardless of fluoroscopy time due to less density over the automatic exposure control cells. We test the hypothesis that air enema reduction of ileocolic intussusception results in lower radiation dose than liquid contrast enema independent of fluoroscopy time. We describe a role for automatic exposure control in this dose difference. We retrospectively evaluated air and liquid intussusception reductions performed on a single digital fluoroscopic unit during a 26-month period. We compared patient age, weight, gender, exam time of day and year, performing radiologist(s), radiographic image acquisitions, grid and magnification use, fluoroscopy time and dose area product. We compared categorical and continuous variables statistically using chi-square and Mann-Whitney U tests, respectively. The mean dose area product was 2.7-fold lower for air enema, 1.3 ± 0.9 dGy·cm 2 , than for liquid, 3.5 ± 2.5 dGy·cm 2 (P<0.005). The mean fluoroscopy time was similar between techniques. The mean dose area product/min was 2.3-fold lower for air, 0.6 ± 0.2 dGy·cm 2 /min, than for liquid, 1.4 ± 0.5 dGy·cm 2 /min (P<0.001). No group differences were identified in other measured dose parameters. Fluoroscopic intussusception reduction using air enema uses less than half the radiation dose of liquid contrast enema. Dose savings are independent of fluoroscopy time and are likely due to automatic exposure control interaction.

  14. Bench-Top Feasibility Testing of a Novel Percutaneous Renal Access Technique: The Laser Direct Alignment Radiation Reduction Technique (DARRT).

    PubMed

    Khater, Nazih; Shen, Jim; Arenas, Javier; Keheila, Mohamed; Alsyouf, Muhannad; Martin, Jacob A; Lightfoot, Michelle A; Li, Roger; Olgin, Gaudencio; Smith, Jason C; Baldwin, D Duane

    2016-11-01

    Traditional techniques for obtaining percutaneous renal access utilize continuous fluoroscopy. In an attempt to minimize radiation exposure, we describe a novel laser direct alignment radiation reduction technique (DARRT) for percutaneous access and test it in a bench-top model. In this randomized-controlled bench-top study, 20 medical personnel obtained renal accesses using both the conventional bullseye technique and the laser DARRT. The primary endpoint was total fluoroscopy time. Secondary endpoints included insertion time, puncture attempts, course corrections, and subjective procedural difficulty. In the laser DARRT, fluoroscopy was used with the C-arm positioned with the laser beam at a 30° angle. The access needle and hub were aligned with the laser beam. Effective caliceal puncture was confirmed with fluoroscopy and direct vision. The Paired samples Wilcoxon signed rank test was used for statistical analysis with significance at p < 0.05. A total of 120 needle placements were recorded. Fluoroscopy time for needle access using the laser DARRT was significantly lower than the bullseye technique in all groups as follows: attendings (7.09 vs 18.51 seconds; p < 0.001), residents (6.55 vs 13.93 seconds; p = 0.001), and medical students (6.69 vs 20.22 seconds; p < 0.001). Students rated the laser DARRT easier to use (2.56 vs 4.89; p < 0.001). No difference was seen in total access time, puncture attempts, or course corrections between techniques. The laser DARRT reduced fluoroscopy time by 63%, compared with the conventional bullseye technique. The least experienced users found the laser DARRT significantly easier to learn. This novel technique is promising and merits additional testing in animal and human models.

  15. Effectiveness of using low rate fluoroscopy to reduce an examiner's radiation dose during lumbar nerve root block.

    PubMed

    Yamane, Kentaro; Kai, Nobuo; Mazaki, Tetsuro; Miyamoto, Tadashi; Matsushita, Tomohiro

    2018-06-13

    Long-term exposure to radiation can lead to gene mutations and increase the risk of cancer. Low rate fluoroscopy has the potential to reduce the radiation exposure for both the examiner and the patient during various fluoroscopic procedures. The purpose of this study was to evaluate the impact of low rate fluoroscopy on reducing an examiner's radiation dose during nerve root block. A total of 101 lumbar nerve root block examinations were performed at our institute during a 6-month period. During the first 3 months, low rate fluoroscopy was performed at 7.5 frames/s (FPS) in 54 examinations, while 47 were performed at 15 FPS during the last 3 months. The examiner wore a torso protector, a neck protector, radiation protection gloves, and radiation protection glasses. Optically stimulated luminescence (OSL) dosimeter badges were placed on both the inside and the outside of each protector. The dosimeters were exchanged every month. Radiation doses (mSv) were measured as the integrated radiation quantity every month from the OSL dosimeters. The effective and equivalent doses for the hands, skin, and eyes were investigated. The mean monthly equivalent doses were significantly lower both inside and outside the hand protector for the 7.5 FPS versus 15 FPS (inside; P = 0.021, outside; P = 0.024). There were no significant differences between the two groups for the mean monthly calculated effective dose for each protector's condition. Radiation exposure was significantly reduced for the skin on the examiner's hand when using low rate fluoroscopy at 7.5 FPS, with no noticeable decrease in image quality or prolonged fluoroscopy time. Copyright © 2018. Published by Elsevier B.V.

  16. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Bargellini, Irene, E-mail: irenebargellini@hotmail.com; Turini, Francesca; Bozzi, Elena

    To assess feasibility of proper hepatic artery catheterization using a 3D model obtained from preprocedural computed tomographic angiography (CTA), fused with real-time fluoroscopy, during transarterial chemoembolization of hepatocellular carcinoma. Twenty consecutive cirrhotic patients with hepatocellular carcinoma undergoing transarterial chemoembolization were prospectively enrolled onto the study. The early arterial phase axial images of the preprocedural CTA were postprocessed on an independent workstation connected to the angiographic system (Innova 4100; GE Healthcare, Milwaukee, WI), obtaining a 3D volume rendering image (VR) that included abdominal aorta, splanchnic arteries, and first and second lumbar vertebrae. The VR image was manually registered to the real-timemore » X-ray fluoroscopy, with the lumbar spine used as the reference. The VR image was then used as guidance to selectively catheterize the proper hepatic artery. The procedure was considered successful when performed with no need for intraarterial contrast injections or angiographic acquisitions. The procedure was successful in 19 (95 %) of 20 patients. In one patient, celiac trunk angiography was required for the presence of a significant ostial stenosis that was underestimated at computed tomography. Time for image reconstruction and registration was <10 min in all cases. The use of preprocedural CTA model with fluoroscopy enables confident and direct catheterization of the proper hepatic artery with no need for preliminary celiac trunk angiography, thus reducing radiation exposure and contrast media administration.« less

  17. Intravascular US-Guided Portal Vein Access: Improved Procedural Metrics during TIPS Creation.

    PubMed

    Gipson, Matthew G; Smith, Mitchell T; Durham, Janette D; Brown, Anthony; Johnson, Thor; Ray, Charles E; Gupta, Rajan K; Kondo, Kimi L; Rochon, Paul J; Ryu, Robert K

    2016-08-01

    To evaluate transjugular intrahepatic portosystemic shunt (TIPS) outcomes and procedure metrics with the use of three different image guidance techniques for portal vein (PV) access during TIPS creation. A retrospective review of consecutive patients who underwent TIPS procedures for a range of indications during a 28-month study period identified a population of 68 patients. This was stratified by PV access techniques: fluoroscopic guidance with or without portography (n = 26), PV marker wire guidance (n = 18), or intravascular ultrasound (US) guidance (n = 24). Procedural outcomes and procedural metrics, including radiation exposure, contrast agent volume used, procedure duration, and PV access time, were analyzed. No differences in demographic or procedural characteristics were found among the three groups. Technical success, technical success of the primary planned approach, hemodynamic success, portosystemic gradient, and procedure-related complications were not significantly different among groups. Fluoroscopy time (P = .003), air kerma (P = .01), contrast agent volume (P = .003), and total procedural time (P = .02) were reduced with intravascular US guidance compared with fluoroscopic guidance. Fluoroscopy time (P = .01) and contrast agent volume (P = .02) were reduced with intravascular US guidance compared with marker wire guidance. Intravascular US guidance of PV access during TIPS creation not only facilitates successful TIPS creation in patients with challenging anatomy, as suggested by previous investigations, but also reduces important procedure metrics including radiation exposure, contrast agent volume, and overall procedure duration compared with fluoroscopically guided TIPS creation. Copyright © 2016 SIR. Published by Elsevier Inc. All rights reserved.

  18. A combination of spatial and recursive temporal filtering for noise reduction when using region of interest (ROI) fluoroscopy for patient dose reduction in image guided vascular interventions with significant anatomical motion

    NASA Astrophysics Data System (ADS)

    Setlur Nagesh, S. V.; Khobragade, P.; Ionita, C.; Bednarek, D. R.; Rudin, S.

    2015-03-01

    Because x-ray based image-guided vascular interventions are minimally invasive they are currently the most preferred method of treating disorders such as stroke, arterial stenosis, and aneurysms; however, the x-ray exposure to the patient during long image-guided interventional procedures could cause harmful effects such as cancer in the long run and even tissue damage in the short term. ROI fluoroscopy reduces patient dose by differentially attenuating the incident x-rays outside the region-of-interest. To reduce the noise in the dose-reduced regions previously recursive temporal filtering was successfully demonstrated for neurovascular interventions. However, in cardiac interventions, anatomical motion is significant and excessive recursive filtering could cause blur. In this work the effects of three noise-reduction schemes, including recursive temporal filtering, spatial mean filtering, and a combination of spatial and recursive temporal filtering, were investigated in a simulated ROI dose-reduced cardiac intervention. First a model to simulate the aortic arch and its movement was built. A coronary stent was used to simulate a bioprosthetic valve used in TAVR procedures and was deployed under dose-reduced ROI fluoroscopy during the simulated heart motion. The images were then retrospectively processed for noise reduction in the periphery, using recursive temporal filtering, spatial filtering and a combination of both. Quantitative metrics for all three noise reduction schemes are calculated and are presented as results. From these it can be concluded that with significant anatomical motion, a combination of spatial and recursive temporal filtering scheme is best suited for reducing the excess quantum noise in the periphery. This new noise-reduction technique in combination with ROI fluoroscopy has the potential for substantial patient-dose savings in cardiac interventions.

  19. Ultrasound-Guided Angioplasty of Dysfunctional Vascular Access for Haemodialysis. The Pros and Cons

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    García-Medina, J., E-mail: josegmedina57@gmail.com; García-Alfonso, J. J., E-mail: juanjozarandieta@gmail.com

    PurposeTo describe the benefits and the disadvantages of angioplasty in dialysis fistulas using only ultrasound guidance.Materials and MethodsThis is a prospective study in 132 failing or non-maturing arteriovenous accesses that underwent 189 ultrasound-guided balloon angioplasties. The technical success was defined as non-use of X-ray fluoroscopy during the procedure.Results127 procedures (67%) were successfully completed without fluoroscopy. Most failures were due to difficulty to traverse aneurismal segments, as well as anastomotic stenoses. Including initial failures, the primary patency rates at 6, 12 months and 2 years were 75 ± 3, 41 ± 3 and 14 ± 2%, respectively.ConclusionEndovascular repair of the dysfunctional vascular access for haemodialysis under ultrasound guidance ismore » feasible and safe in roughly two-thirds of cases.« less

  20. Lightweight bilayer barium sulfate-bismuth oxide composite thyroid collars for superior radiation protection in fluoroscopy-guided interventions: a prospective randomized controlled trial.

    PubMed

    Uthoff, Heiko; Benenati, Matthew J; Katzen, Barry T; Peña, Constantino; Gandhi, Ripal; Staub, Daniel; Schernthaner, Melanie

    2014-02-01

    To test whether newer bilayer barium sulfate-bismuth oxide composite (XPF) thyroid collars (TCs) provide superior radiation protection and comfort during fluoroscopy-guided interventions compared with standard 0.5-mm lead-equivalent TCs. Institutional review board approval and written informed consent were obtained for this HIPAA-compliant study, and 144 fluoroscopy-guided vascular interventions were included at one center between October 2011 and July 2012, with up to two operators randomly assigned to wear XPF (n = 135) or standard 0.5-mm lead-equivalent (n = 121) TCs. Radiation doses were measured by using dosimeters placed outside and underneath the TCs. Wearing comfort was assessed at the end of each procedure on a visual analog scale (0-100, with 100 indicating optimal comfort). Adjusted differences in comfort and radiation dose reductions were calculated by using a mixed logistic regression model and the common method of inverse variance weighting, respectively. Patient (height, weight, and body mass index) and procedure (type and duration of intervention, operator, fluoroscopy time, dose-area product, and air kerma) data did not differ between the XPF and standard groups. Comfort was assessed in all 256 measurements. On average, the XPF TCs were 47.6% lighter than the standard TCs (mean weight ± standard deviation, 133 g ± 14 vs 254 g ± 44; P < .001) and had a significantly higher likelihood of a high level of comfort (visual analog scale >90; odds ratio, 7.6; 95% confidence interval: 3.0, 19.2; P < .001). Radiation dose reduction provided by the TCs was analyzed in 117 data sets (60 in the XPF group, 57 in the standard group). The mean radiation dose reductions (ie, radiation protection) provided by XPF and standard TCs were 90.7% and 72.4%, with an adjusted mean difference of 17.9% (95% confidence interval: 7.7%, 28.1%; P < .001) favoring XPF. XPF TCs are a lightweight alternative to standard 0.5-mm lead-equivalent TCs and provide superior radiation protection during fluoroscopy-guided interventions. © RSNA, 2013.

  1. Impact of Anatomical, Procedural, and Operator Skill Factors on the Success and Duration of Fluoroscopy-Guided Transjugular Intrahepatic Portosystemic Shunt

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Marquardt, Steffen, E-mail: marquardt.steffen@mh-hannover.de; Rodt, Thomas, E-mail: rodt.thomas@mh-hannover.de; Rosenthal, Herbert, E-mail: rosenthal.herbert@mh-hannover.de

    PurposeTo assess the impact of anatomical, procedural, and operator skill factors on the success and duration of fluoroscopy-guided transjugular intrahepatic portoystemic shunt following standard operating procedure (SOP).Material and MethodsDuring a 32-month period, 102 patients underwent transjugular intrahepatic portosystemic shunt creation (TIPS) by two interventional radiologists (IR) following our institutional SOP based on fluoroscopy guidance. Both demographic and procedural data were assessed. The duration of the intervention (D{sub Int}) and of the portal vein puncture (D{sub Punct}) was analyzed depending on the skill level of the IR as well as the anatomic or procedural factors.ResultsIn 99 of the 102 patients, successfulmore » TIPS without peri-procedural complications was performed. The mean D{sub Int} (IR1: 77 min; IR2: 51 min, P < 0.005) and the mean D{sub Punct} (IR1: 19 min; IR2: 13 min, P < 0.005) were significantly higher in TIPS performed by IR1 (with 2 years of clinical experience performing TIPS, n = 38) than by IR2 (>10 years of clinical experience performing TIPS, n = 61), (P < 0.005 both, Mann–Whitney U test). D{sub Int} showed a higher correlation with D{sub Punct} for IR2 (R{sup 2} = 0.63) than for IR1 (R{sup 2} = 0.13). There was no significant difference in the D{sub Punct} for both IRs with regard to the success of the wedged portography (P = 0.90), diameter of the portal vein (P = 0.60), central right portal vein length (P = 0.49), or liver function (MELD-Score before the TIPS procedure; P = 0.14).ConclusionTIPS following SOP is safe, fast, and reliable. The only significant factor for shorter D{sub Punct} and D{sub Int} was the clinical experience of the IR. Anatomic variability, successful portography, or liver function did not alter the duration or technical success of TIPS.« less

  2. SU-G-IeP3-13: Real-Time Patient and Staff Dose Monitoring in Fluoroscopy Guided Interventions

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Vergoossen, L; Sailer, A; Paulis, L

    Purpose: Interventional radiology procedures involve the use of X-rays, which can pose a large radiation burden on both patients and staff. Although some reports on radiation dose are available, most studies focus on limited types of procedures and only report patient dose. In our cathlabs a dedicated real-time patient and staff monitoring system was installed in November 2015. The aim of this study was to investigate the patient and staff dose exposure for different types of interventions. Methods: Radiologists involved in fluoroscopy guided interventional radiology procedures wore personal dose meters (PDM, DoseAware, Philips) on their lead-apron that measured the personalmore » dose equivalent Hp(10), a measure for the effective dose (E). Furthermore, reference PDMs were installed in the C-arms of the fluoroscopy system (Allura XPer, Philips). Patient dose-area-product (DAP) and PDM doses were retrieved from the monitoring system (DoseWise, Philips) for each procedure. A total of 399 procedures performed between November 2015 and February 2016 were analyzed with respect to the type of intervention. Interventions were grouped by anatomy and radiologist position. Results: The mean DAP for the different types of interventions ranged from 2.86±2.96 Gycm{sup 2} (percutaneous gastrostomy) to 147±178 Gycm{sup 2} (aortic repair procedures). The radiologist dose (E) ranged from 5.39±7.38 µSv (cerebral interventions) to 84.7±106 µSv (abdominal interventions) and strongly correlated with DAP (R{sup 2}=0.83). The E normalized to DAP showed that the relative radiologist dose was higher for interventions in larger body parts (e.g. abdomen) compared to smaller body parts (e.g. head). Conclusion: Using a real-time dose monitoring system we were able to assess the staff and patient dose revealing that the relative staff dose strongly depended on the type of procedure and patient anatomy. This could be explained by the position of the radiologist with respect to the patient and X-ray tube. To facilitate this study L Vergoossen received a scholarship from Philips Medical Systems.« less

  3. A novel technique for ventriculoperitoneal shunting by flat panel detector CT-guided real-time fluoroscopy

    PubMed Central

    Kobayashi, Shinya; Ishikawa, Tatsuya; Mutoh, Tatsushi; Hikichi, Kentaro; Suzuki, Akifumi

    2012-01-01

    Background: Surgical placement of a ventriculoperitoneal shunt (VPS) is the main strategy to manage hydrocephalus. However, the failure rate associated with placement of ventricular catheters remains high. Methods: A hybrid operating room, equipped with a flat-panel detector digital subtraction angiography system containing C-arm cone-beam computed tomography (CB-CT) imaging, has recently been developed and utilized to assist neurosurgical procedures. We have developed a novel technique using intraoperative fluoroscopy and a C-arm CB-CT system to facilitate accurate placement of a VPS. Results: Using this novel technique, 39 consecutive ventricular catheters were placed accurately, and no ventricular catheter failures were experienced during the follow-up period. Only two patients experienced obstruction of the VPS, both of which occurred in the extracranial portion of the shunt system. Conclusion: Surgical placement of a VPS assisted by flat panel detector CT-guided real-time fluoroscopy enabled accurate placement of ventricular catheters and was associated with a decreased need for shunt revision. PMID:23226605

  4. Towards image-guided atrial septal defect repair: an ex vivo analysis

    NASA Astrophysics Data System (ADS)

    Kwartowitz, David M.; Mefleh, Fuad N.; Baker, George H.

    2012-02-01

    The use of medical images in the operating room for navigation and planning is well established in many clinical disciplines. In cardiology, the use of fluoroscopy for the placement of catheters within the heart has become the standard of care. While fluoroscopy provides a live video sequence with the current location, it poses risks the patient and clinician through exposure to radiation. Radiation dose is cumulative and thus children are at even greater risk from exposure. To reduce the use of radiation, and improve surgical technique we have begun development of an image-guided navigation system, which can deliver therapeutic devices via catheter. In this work we have demonstrated the intrinsic properties of our imaging system, which have led to the development of a phantom emulating a childs heart with an ASD. Further investigation into the use of this information, in a series of mock clinical experiments, will be performed to design procedures for inserting devices into the heart while minimizing fluoroscopy use.

  5. [Evaluation of an Experimental Production Wireless Dose Monitoring System for Radiation Exposure Management of Medical Staff].

    PubMed

    Fujibuchi, Toshioh; Murazaki, Hiroo; Kuramoto, Taku; Umedzu, Yoshiyuki; Ishigaki, Yung

    2015-08-01

    Because of the more advanced and more complex procedures in interventional radiology, longer treatment times have become necessary. Therefore, it is important to determine the exposure doses received by operators and patients. The aim of our study was to evaluate an experimental production wireless dose monitoring system for pulse radiation in diagnostic X-ray. The energy, dose rate, and pulse fluoroscopy dependence were evaluated as the basic characteristics of this system for diagnostic X-ray using a fully digital fluoroscopy system. The error of 1 cm dose equivalent rate was less than 15% from 35.1 keV to 43.2 keV with energy correction using metal filter. It was possible to accurately measure the dose rate dependence of this system, which was highly linear until 100 μSv/h. This system showed a constant response to the pulse fluoroscopy. This system will become useful wireless dosimeter for the individual exposure management by improving the high dose rate and the energy characteristics.

  6. A motion-compensated image filter for low-dose fluoroscopy in a real-time tumor-tracking radiotherapy system

    PubMed Central

    Miyamoto, Naoki; Ishikawa, Masayori; Sutherland, Kenneth; Suzuki, Ryusuke; Matsuura, Taeko; Toramatsu, Chie; Takao, Seishin; Nihongi, Hideaki; Shimizu, Shinichi; Umegaki, Kikuo; Shirato, Hiroki

    2015-01-01

    In the real-time tumor-tracking radiotherapy system, a surrogate fiducial marker inserted in or near the tumor is detected by fluoroscopy to realize respiratory-gated radiotherapy. The imaging dose caused by fluoroscopy should be minimized. In this work, an image processing technique is proposed for tracing a moving marker in low-dose imaging. The proposed tracking technique is a combination of a motion-compensated recursive filter and template pattern matching. The proposed image filter can reduce motion artifacts resulting from the recursive process based on the determination of the region of interest for the next frame according to the current marker position in the fluoroscopic images. The effectiveness of the proposed technique and the expected clinical benefit were examined by phantom experimental studies with actual tumor trajectories generated from clinical patient data. It was demonstrated that the marker motion could be traced in low-dose imaging by applying the proposed algorithm with acceptable registration error and high pattern recognition score in all trajectories, although some trajectories were not able to be tracked with the conventional spatial filters or without image filters. The positional accuracy is expected to be kept within ±2 mm. The total computation time required to determine the marker position is a few milliseconds. The proposed image processing technique is applicable for imaging dose reduction. PMID:25129556

  7. The effect of bench model fidelity on fluoroscopy-guided transforaminal epidural injection training: a randomized control study.

    PubMed

    Gonzalez-Cota, Alan; Chiravuri, Srinivas; Stansfield, R Brent; Brummett, Chad M; Hamstra, Stanley J

    2013-01-01

    The purpose of this study was to determine whether high-fidelity simulators provide greater benefit than low-fidelity models in training fluoroscopy-guided transforaminal epidural injection. This educational study was a single-center, prospective, randomized 3-arm pretest-posttest design with a control arm. Eighteen anesthesia and physical medicine and rehabilitation residents were instructed how to perform a fluoroscopy-guided transforaminal epidural injection and assessed by experts on a reusable injectable phantom cadaver. The high- and low-fidelity groups received 30 minutes of supervised hands-on practice according to group assignment, and the control group received 30 minutes of didactic instruction from an expert. We found no differences at posttest between the high- and low-fidelity groups on global ratings of performance (P = 0.17) or checklist scores (P = 0.81). Participants who received either form of hands-on training significantly outperformed the control group on both the global rating of performance (control vs low-fidelity, P = 0.0048; control vs high-fidelity, P = 0.0047) and the checklist (control vs low-fidelity, P = 0.0047; control vs high-fidelity, P = 0.0047). Training an epidural procedure using a low-fidelity model may be equally effective as training on a high-fidelity model. These results are consistent with previous research on a variety of interventional procedures and further demonstrate the potential impact of simple, low-fidelity training models.

  8. Imaging Techniques in Percutaneous Cardiac Structural Interventions: Atrial Septal Defect Closure and Left Atrial Appendage Occlusion.

    PubMed

    Rodríguez Fernández, Antonio; Bethencourt González, Armando

    2016-08-01

    Because of advances in cardiac structural interventional procedures, imaging techniques are playing an increasingly important role. Imaging studies show sufficient anatomic detail of the heart structure to achieve an excellent outcome in interventional procedures. Up to 98% of atrial septal defects at the ostium secundum can be closed successfully with a percutaneous procedure. Candidates for this type of procedure can be identified through a systematic assessment of atrial septum anatomy, locating and measuring the size and shape of all defects, their rims, and the degree and direction of shunting. Three dimensional echocardiography has significantly improved anatomic assessments and the end result itself. In the future, when combined with other imaging techniques such as cardiac computed tomography and fluoroscopy, 3-dimensional echocardiography will be particularly useful for procedure guidance. Percutaneous closure of the left atrial appendage offers an alternative for treating patients with atrial fibrillation and contraindication for oral anticoagulants. In the future, the clinical focus may well turn to stroke prevention in selected patients. Percutaneous closure is effective and safe; device implantation is successful in 94% to 99% of procedures. However, the procedure requires an experienced cardiac structural interventional team. At present, 3-dimensional echocardiography is the most appropriate imaging technique to assess anatomy suitability, select device type and size, guide the procedure alongside fluoroscopy, and to follow-up the patient afterwards. Copyright © 2016 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved.

  9. Patient dose in interventional radiology: a multicentre study of the most frequent procedures in France.

    PubMed

    Etard, Cécile; Bigand, Emeline; Salvat, Cécile; Vidal, Vincent; Beregi, Jean Paul; Hornbeck, Amaury; Greffier, Joël

    2017-10-01

    A national retrospective survey on patient doses was performed by the French Society of Medical physicists to assess reference levels (RLs) in interventional radiology as required by the European Directive 2013/59/Euratom. Fifteen interventional procedures in neuroradiology, vascular radiology and osteoarticular procedures were analysed. Kerma area product (KAP), fluoroscopy time (FT), reference air kerma and number of images were recorded for 10 to 30 patients per procedure. RLs were calculated as the 3rd quartiles of the distributions. Results on 4600 procedures from 36 departments confirmed the large variability in patient dose for the same procedure. RLs were proposed for the four dosimetric estimators and the 15 procedures. RLs in terms of KAP and FT were 90 Gm.cm 2 and 11 mins for cerebral angiography, 35 Gy.cm 2 and 16 mins for biliary drainage, 75 Gy.cm 2 and 6 mins for lower limbs arteriography and 70 Gy.cm 2 and 11 mins for vertebroplasty. For these four procedures, RLs were defined according to the complexity of the procedure. For all the procedures, the results were lower than most of those already published. This study reports RLs in interventional radiology based on a national survey. Continual evolution of practices and technologies requires regular updates of RLs. • Delivered dose in interventional radiology depends on procedure, practice and patient. • National RLs are proposed for 15 interventional procedures. • Reference levels (RLs) are useful to benchmark practices and optimize protocols. • RLs are proposed for kerma area product, air kerma, fluoroscopy time and number of images. • RLs should be adapted to the procedure complexity and updated regularly.

  10. Balloon-augmented Onyx embolization of cerebral arteriovenous malformations using a dual-lumen balloon: a multicenter experience.

    PubMed

    Spiotta, Alejandro M; James, Robert F; Lowe, Stephen R; Vargas, Jan; Turk, Aquilla S; Chaudry, M Imran; Bhalla, Tarun; Janjua, Rashid M; Delaney, John J; Quintero-Wolfe, Stacey; Turner, Raymond D

    2015-10-01

    Conventional Onyx embolization of cerebral arteriovenous malformations (AVMs) requires lengthy procedure and fluoroscopy times to form an adequate 'proximal plug' which allows forward nidal penetration while preventing reflux and non-targeted embolization. We review our experience with balloon-augmented Onyx embolization of cerebral AVMs using a dual-lumen balloon catheter technique designed to minimize these challenges. Retrospectively acquired data for all balloon-augmented cerebral AVM embolizations performed between 2011 and 2014 were obtained from four tertiary care centers. For each procedure, at least one Scepter C balloon catheter was advanced into the AVM arterial pedicle of interest and Onyx embolization was performed through the inner lumen after balloon inflation via the outer lumen. Twenty patients underwent embolization with the balloon-augmented technique over 24 discreet treatment episodes. There were 37 total arterial pedicles embolized with the balloon-augmented technique, a mean of 1.9 per patient (range 1-5). The treated AVMs were heterogeneous in their location and size (mean 3.3±1.6 cm). Mean fluoroscopy time for each procedure was 48±26 min (28 min per embolized pedicle). Two Scepter C balloon catheter-related complications (8.3% of embolization sessions, 5.4% of pedicles embolized) were observed: an intraprocedural rupture of a feeding pedicle and fracture and retention of a catheter fragment. This multicenter experience represents the largest reported series of balloon-augmented Onyx embolization of cerebral AVMs. The technique appears safe and effective in the treatment of AVMs, allowing more efficient and controlled injection of Onyx with a decreased risk of reflux and decreased fluoroscopy times. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions.

  11. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sailer, Anna M., E-mail: anni.sailer@mumc.nl; Haan, Michiel W. de, E-mail: m.de.haan@mumc.nl; Graaf, Rick de, E-mail: r.de.graaf@mumc.nl

    PurposeThis study was designed to evaluate the feasibility of endovascular guidance by means of live fluoroscopy fusion with magnetic resonance angiography (MRA) and computed tomography angiography (CTA).MethodsFusion guidance was evaluated in 20 endovascular peripheral artery interventions in 17 patients. Fifteen patients had received preinterventional diagnostic MRA and two patients had undergone CTA. Time for fluoroscopy with MRA/CTA coregistration was recorded. Feasibility of fusion guidance was evaluated according to the following criteria: for every procedure the executing interventional radiologists recorded whether 3D road-mapping provided added value (yes vs. no) and whether PTA and/or stenting could be performed relying on the fusionmore » road-map without need for diagnostic contrast-enhanced angiogram series (CEAS) (yes vs. no). Precision of the fusion road-map was evaluated by recording maximum differences between the position of the vasculature on the virtual CTA/MRA images and conventional angiography.ResultsAverage time needed for image coregistration was 5 ± 2 min. Three-dimensional road-map added value was experienced in 15 procedures in 12 patients. In half of the patients (8/17), intervention was performed relying on the fusion road-map only, without diagnostic CEAS. In two patients, MRA roadmap showed a false-positive lesion. Excluding three patients with inordinate movements, mean difference in position of vasculature on angiography and MRA/CTA road-map was 1.86 ± 0.95 mm, implying that approximately 95 % of differences were between 0 and 3.72 mm (2 ± 1.96 standard deviation).ConclusionsFluoroscopy with MRA/CTA fusion guidance for peripheral artery interventions is feasible. By reducing the number of CEAS, this technology may contribute to enhance procedural safety.« less

  12. A posterior approach to cervical nerve root block and pulsed radiofrequency treatment for cervical radicular pain: a retrospective study.

    PubMed

    Xiao, Lizu; Li, Jie; Li, Disen; Yan, Dong; Yang, Jun; Wang, Daniel; Cheng, Jianguo

    2015-09-01

    Catastrophic complications have been reported for selective cervical nerve root block (SCNRB) or pulsed radiofrequency (PRF) via an anterolateral transforaminal approach. A posterior approach to these procedures under computed tomography guidance has been reported. Here, we report the clinical outcomes of 42 patients with chronic cervical radicular pain (CCRP) treated with a combination of SCNRB and PRF through a posterior approach under fluoroscopy guidance. We retrospectively reviewed the clinical outcomes of 42 consecutive patients with CCRP who received a combination of SCNRB and PRF through a posterior approach under fluoroscopy guidance. The thresholds of electrical stimulation and imaging of the nerve roots after contrast injection were used to evaluate the accuracy of needle placement. The numeric rating scale was used to measure the pain and numbness levels as primary clinical outcomes, which were evaluate in scheduled follow-up visits of up to 3 months. A total of 53 procedures were performed on 42 patients at the levels of C5-C8. All patients reported concordant paresthesia in response to electrical stimulation. The average sensory and motor thresholds of stimulation were 0.28 ± 0.14 and 0.36 ± 0.14 V, respectively. Injection of nonionic contrast resulted in excellent spread along the target nerve root in large majority of the procedures. The numeric rating scale scores for both pain and numbness improved significantly at 1 day, 1 week, and 1 and 3 months after the treatment. No serious adverse effects were observed in any of the patients. The posterior approach to combined SCNRB and PRF under fluoroscopy guidance appears to be safe and efficacious in the management of CCRP. Copyright © 2015 Elsevier Inc. All rights reserved.

  13. Sensor-based electromagnetic navigation to facilitate implantation of left ventricular leads in cardiac resynchronization therapy.

    PubMed

    Döring, Michael; Sommer, Philipp; Rolf, Sascha; Lucas, Johannes; Breithardt, Ole A; Hindricks, Gerhard; Richter, Sergio

    2015-02-01

    Implantation of cardiac resynchronization therapy (CRT) devices can be challenging, time consuming, and fluoroscopy intense. To facilitate placement of left ventricular (LV) leads, a novel electromagnetic navigation system (MediGuide™, St. Jude Medical, St. Paul, MN, USA) has been developed, displaying real-time 3-D location of sensor-embedded delivery tools superimposed on prerecorded X-ray cine-loops of coronary sinus venograms. We report our experience and advanced progress in the use of this new electromagnetic tracking system to guide LV lead implantation. Between January 2012 and December 2013, 71 consecutive patients (69 ± 9 years, 76% male) were implanted with a CRT device using the new electromagnetic tracking system. Demographics, procedural data, and periprocedural adverse events were gathered. The impact of the operator's experience, optimized workflow, and improved software technology on procedural data were analyzed. LV lead implantation was successfully achieved in all patients without severe adverse events. Total procedure time measured 87 ± 37 minutes and the median total fluoroscopy time (skin-to-skin) was 4.9 (2.5-7.8) minutes with a median dose-area-product of 476 (260-1056) cGy*cm(2) . An additional comparison with conventional CRT device implantations showed a significant reduction in fluoroscopy time from 8.0 (5.8; 11.5) to 4.5 (2.8; 7.3) minutes (P = 0.016) and radiation dose from 603 (330; 969) to 338 (176; 680) cGy*cm(2) , respectively (P = 0.044 ). Use of the new navigation system enables safe and successful LV lead placement with improved orientation and significantly reduced radiation exposure during CRT implantation. © 2014 Wiley Periodicals, Inc.

  14. A novel technique for transcatheter patent ductus arteriosus closure in extremely preterm infants using commercially available technology.

    PubMed

    Zahn, Evan M; Nevin, Phillip; Simmons, Charles; Garg, Ruchira

    2015-02-01

    To describe a new technique for transcatheter patent ductus arteriosus (PDA) closure in extremely preterm infants using commercially available technology. PDA in premature neonates continues to be a significant clinical problem contributing importantly to both morbidity and mortality. Surgical ligation and medical therapy both have their drawbacks. Hospital records and catheterization reports of all premature neonates (< 32 weeks gestation) who underwent transcatheter PDA closure between March 2013 and February 2014 were reviewed. Particular attention was paid to procedural details, complications, and short and mid-term outcomes. Six premature infants born at gestational ages ranging between 26 and 31 weeks (median, 26 weeks) underwent attempted transcatheter PDA closure using the Amplatzer Vascular Plug II (AVP II). Median age and weight was 21.5 days (16-80 days) and 1,180 g (870-2,240 g), respectively. Fluoroscopy and echocardiography were used to guide device. Contrast angiography was not used in any patient. Complete closure was achieved in all patients with no major procedural complications. Median fluoroscopy and procedural times were 9.4 (0-19.5) and 51.5 (33-87) min, respectively. All patients were alive at the time of this report. There were no instances of device migration, left pulmonary artery (LPA), or aortic coarctation. This preliminary study demonstrates that transcatheter PDA closure can be successfully performed in extremely preterm neonates using currently available technology with a high success rate and a low incidence of complications. This report also describes a novel transvenous approach using a combination of echocardiography and judicious use of fluoroscopy to avoid arterial access in this fragile patient population. © 2014 Wiley Periodicals, Inc. © 2014 Wiley Periodicals, Inc.

  15. New image-processing and noise-reduction software reduces radiation dose during complex endovascular procedures.

    PubMed

    Kirkwood, Melissa L; Guild, Jeffrey B; Arbique, Gary M; Tsai, Shirling; Modrall, J Gregory; Anderson, Jon A; Rectenwald, John; Timaran, Carlos

    2016-11-01

    A new proprietary image-processing system known as AlluraClarity, developed by Philips Healthcare (Best, The Netherlands) for radiation-based interventional procedures, claims to lower radiation dose while preserving image quality using noise-reduction algorithms. This study determined whether the surgeon and patient radiation dose during complex endovascular procedures (CEPs) is decreased after the implementation of this new operating system. Radiation dose to operators, procedure type, reference air kerma, kerma area product, and patient body mass index were recorded during CEPs on two Philips Allura FD 20 fluoroscopy systems with and without Clarity. Operator dose during CEPs was measured using optically stimulable, luminescent nanoDot (Landauer Inc, Glenwood, Ill) detectors placed outside the lead apron at the left upper chest position. nanoDots were read using a microStar ii (Landauer Inc) medical dosimetry system. For the CEPs in the Clarity group, the radiation dose to surgeons was also measured by the DoseAware (Philips Healthcare) personal dosimetry system. Side-by-side measurements of DoseAware and nanoDots allowed for cross-calibration between systems. Operator effective dose was determined using a modified Niklason algorithm. To control for patient size and case complexity, the average fluoroscopy dose rate and the dose per radiographic frame were adjusted for body mass index differences and then compared between the groups with and without Clarity by procedure. Additional factors, for example, physician practice patterns, that may have affected operator dose were inferred by comparing the ratio of the operator dose to procedural kerma area product with and without Clarity. A one-sided Wilcoxon rank sum test was used to compare groups for radiation doses, reference air kermas, and operating practices for each procedure type. The analysis included 234 CEPs; 95 performed without Clarity and 139 with Clarity. Practice patterns of operators during procedures with and without Clarity were not significantly different. For all cases, procedure radiation dose to the patient and the primary and assistant operators were significantly decreased in the Clarity group by 60% compared with the non-Clarity group. By procedure type, fluorography dose rates decreased from 44% for fenestrated endovascular repair and up to 70% with lower extremity interventions. Fluoroscopy dose rates also significantly decreased, from about 37% to 47%, depending on procedure type. The AlluraClarity system reduces the patient and primary operator's radiation dose by more than half during CEPs. This feature appears to be an effective tool in lowering the radiation dose while maintaining image quality. Copyright © 2016 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

  16. A national patient dose survey and setting of reference levels for interventional radiology in Bulgaria.

    PubMed

    Zotova, R; Vassileva, J; Hristova, J; Pirinen, M; Järvinen, H

    2012-06-01

    A national study on patient dose values in interventional radiology and cardiology was performed in order to assess current practice in Bulgaria, to estimate the typical patient doses and to propose reference levels for the most common procedures. Fifteen units and more than 1,000 cases were included. Average values of the measured parameters for three procedures-coronary angiography (CA), combined procedure (CA + PCI) and lower limb arteriography (LLA)--were compared with data published in the literature. Substantial variations were observed in equipment and procedure protocols used. This resulted in variations in patient dose: air-kerma area product ranges were 4-339, 6-1,003 and 0.2-288 Gy cm(2) for CA, CA + PCI and LLA respectively. Reference levels for air kerma-area product were proposed: 40 Gy cm(2) for CA, 140 Gy cm(2) for CA + PCI and 45 Gy cm(2) for LLA. Auxiliary reference intervals were proposed for other dose-related parameters: fluoroscopy time, number of images and entrance surface air kerma rate in fluoroscopy and cine mode. There is an apparent necessity for improvement in the classification of peripheral procedures and for standardisation of the protocols applied. It is important that patient doses are routinely recorded and compared with reference levels. • Patient doses in interventional radiology are high and vary greatly • Better standardisation of procedures and techniques is needed to improve practice • Dose reference levels for most common procedures are proposed.

  17. MO-F-16A-04: Case Study: Estimation of Peak Skin Dose Following a Physician Reported “High Dose” Case and Sentinel Event Considerations

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Supanich, M; Chu, J; Wehmeyer, A

    2014-06-15

    Purpose: This work offers as a teaching example a reported high dose fluoroscopy case and the workflow the institution followed to self-report a radiation overdose sentinel event to the Joint Commission. Methods: Following the completion of a clinical case in a hybrid OR room with a reported air kerma of >18 Gy at the Interventional Reference Point (IRP) the physicians involved in the case referred study to the institution's Radiation Safety Committee (RSC) for review. The RSC assigned a Diagnostic Medical Physicist (DMP) to estimate the patient's Peak Skin Dose (PSD) and analyze the case. Following the DMP's analysis andmore » estimate of a PSD of >15 Gy the institution's adverse event committee was convened to discuss the case and to self-report the case as a radiation overdose sentinel event to the Joint Commission. The committee assigned a subgroup to perform the root cause analysis and develop institutional responses to the event. Results: The self-reporting of the sentinel event and the associated root cause analysis resulted in several institutional action items that are designed to improve process and safety. A formal reporting and analysis mechanism was adopted to review fluoroscopy cases with air kerma greater than 6 Gy at the IRP. An improved and formalized radiation safety training program for physicians using fluoroscopy equipment was implemented. Additionally efforts already under way to monitor radiation exposure in the Radiology department were expanded to include all fluoroscopy equipment capable of automated dose reporting. Conclusion: The adverse event review process and the root cause analysis following the self-reporting of the sentinel event resulted in policies and procedures that are expected to improve the quality and safe usage of fluoroscopy throughout the institution.« less

  18. Predictors of radiation exposure to providers during percutaneous nephrolithotomy

    PubMed Central

    Wenzler, David L.; Abbott, Joel E.; Su, Jeannie J.; Shi, William; Slater, Richard; Miller, Daniel; Siemens, Michelle J.; Sur, Roger L.

    2017-01-01

    Background: Limited studies have reported on radiation risks of increased ionizing radiation exposure to medical personnel in the urologic community. Fluoroscopy is readily used in many urologic surgical procedures. The aim of this study was to determine radiation exposure to all operating room personnel during percutaneous nephrolithotomy (PNL), commonly performed for large renal or complex stones. Materials and Methods: We prospectively collected personnel exposure data for all PNL cases at two academic institutions. This was collected using the Instadose™ dosimeter and reported both continuously and categorically as high and low dose using a 10 mrem dose threshold, the approximate amount of radiation received from one single chest X-ray. Predictors of increased radiation exposure were determined using multivariate analysis. Results: A total of 91 PNL cases in 66 patients were reviewed. Median surgery duration and fluoroscopy time were 142 (38–368) min and 263 (19–1809) sec, respectively. Median attending urologist, urology resident, anesthesia, and nurse radiation exposure per case was 4 (0–111), 4 (0–21), 0 (0–5), and 0 (0–5) mrem, respectively. On univariate analysis, stone area, partial or staghorn calculi, surgery duration, and fluoroscopy time were associated with high attending urologist and resident radiation exposure. Preexisting access that was utilized was negatively associated with resident radiation exposure. However, on multivariate analysis, only fluoroscopy duration remained significant for attending urologist radiation exposure. Conclusion: Increased stone burden, partial or staghorn calculi, surgery and fluoroscopy duration, and absence of preexisting access were associated with high provider radiation exposure. Radiation safety awareness is essential to minimize exposure and to protect the patient and all providers from potential radiation injury. PMID:28216931

  19. Pulmonary Masses: Initial Results of Cone-beam CT Guidance with Needle Planning Software for Percutaneous Lung Biopsy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Braak, Sicco J., E-mail: sjbraak@gmail.com; Herder, Gerarda J. M., E-mail: j.herder@antoniusziekenhuis.nl; Heesewijk, Johannes P. M. van, E-mail: j.heesewijk@antoniusziekenhuis.nl

    2012-12-15

    Purpose: To evaluate the outcome of percutaneous lung biopsy (PLB) findings using cone-beam computed tomographic (CT) guidance (CBCT guidance) and compared to conventional biopsy guidance techniques. Methods: CBCT guidance is a stereotactic technique for needle interventions, combining 3D soft-tissue cone-beam CT, needle planning software, and real-time fluoroscopy. Between March 2007 and August 2010, we performed 84 Tru-Cut PLBs, where bronchoscopy did not provide histopathologic diagnosis. Mean patient age was 64.6 (range 24-85) years; 57 patients were men, and 25 were women. Records were prospectively collected for calculating sensitivity, specificity, positive predictive value, negative predictive value, and accuracy. We also registeredmore » fluoroscopy time, room time, interventional time, dose-area product (DAP), and complications. Procedures were divided into subgroups (e.g., location, size, operator). Results: Mean lesion diameter was 32.5 (range 3.0-93.0) mm, and the mean number of samples per biopsy procedure was 3.2 (range 1-7). Mean fluoroscopy time was 161 (range 104-551) s, room time was 34 (range 15-79) min, mean DAP value was 25.9 (range 3.9-80.5) Gy{center_dot}cm{sup -2}, and interventional time was 18 (range 5-65) min. Of 84 lesions, 70 were malignant (83.3%) and 14 were benign (16.7%). Seven (8.3%) of the biopsy samples were nondiagnostic. All nondiagnostic biopsied lesions proved to be malignant during surgical resection. The outcome for sensitivity, specificity, positive predictive value, negative predictive value, and accuracy was 90% (95% confidence interval [CI] 86-96), 100% (95% CI 82-100), 100% (95% CI 96-100), 66.7% (95% CI 55-83), and 91.7% (95% CI 86-96), respectively. Sixteen patients (19%) had minor and 2 (2.4%) had major complications. Conclusion: CBCT guidance is an effective method for PLB, with results comparable to CT/CT fluoroscopy guidance.« less

  20. 2D projection-based software application for mobile C-arms optimises wire placement in the proximal femur - An experimental study.

    PubMed

    Swartman, B; Frere, D; Wei, W; Schnetzke, M; Beisemann, N; Keil, H; Franke, J; Grützner, P A; Vetter, S Y

    2017-10-01

    A new software application can be used without fixed reference markers or a registration process in wire placement. The aim was to compare placement of Kirschner wires (K-wires) into the proximal femur with the software application versus the conventional method without guiding. As study hypothesis, we assumed less placement attempts, shorter procedure time and shorter fluoroscopy time using the software. The same precision inside a proximal femur bone model using the software application was premised. The software detects a K-wire within the 2D fluoroscopic image. By evaluating its direction and tip location, it superimposes a trajectory on the image, visualizing the intended direction of the K-wire. The K-wire was positioned in 20 artificial bones with the use of software by one surgeon; 20 bones served as conventional controls. A brass thumb tack was placed into the femoral head and its tip targeted with the wire. Number of placement attempts, duration of the procedure, duration of fluoroscopy time and distance to the target in a postoperative 3D scan were recorded. Compared with the conventional method, use of the application showed fewer attempts for optimal wire placement (p=0.026), shorter duration of surgery (p=0.004), shorter fluoroscopy time (p=0.024) and higher precision (p=0.018). Final wire position was achieved in the first attempt in 17 out of 20 cases with the software and in 9 out of 20 cases with the conventional method. The study hypothesis was confirmed. The new application optimised the process of K-wire placement in the proximal femur in an artificial bone model while also improving precision. Benefits lie especially in the reduction of placement attempts and reduction of fluoroscopy time under the aspect of radiation protection. The software runs on a conventional image intensifier and can therefore be easily integrated into the daily surgical routine. Copyright © 2017 Elsevier Ltd. All rights reserved.

  1. Image Guidance Technologies for Interventional Pain Procedures: Ultrasound, Fluoroscopy, and CT.

    PubMed

    Wang, Dajie

    2018-01-26

    Chronic pain is a common medical condition. Patients who suffer uncontrolled chronic pain may require interventions including spinal injections and various nerve blocks. Interventional procedures have evolved and improved over time since epidural injection was first introduced for low back pain and sciatica in 1901. One of the major contributors in the improvement of these interventions is the advancement of imaging guidance technologies. The utilization of image guidance has dramatically improved the accuracy and safety of these interventions. The first image guidance technology adopted by pain specialists was fluoroscopy. This was followed by CT and ultrasound. Fluoroscopy can be used to visualize bony structures of the spine. It is still the most commonly used guidance technology in spinal injections. In the recent years, ultrasound guidance has been increasingly adopted by interventionists to perform various injections. Because its ability to visualize soft tissue, vessels, and nerves, this guidance technology appears to be a better option than fluoroscopy for interventions including SGB and celiac plexus blocks, when visualization of the vessels may prevent intravascular injection. The current evidence indicates the efficacies of these interventions are similar between ultrasound guidance and fluoroscopy guidance for SGB and celiac plexus blocks. For facet injections and interlaminar epidural steroid injections, it is important to visualize bony structures in order to perform these procedures accurately and safely. It is worth noting that facet joint injections can be done under ultrasound guidance with equivalent efficacy to fluoroscopic guidance. However, obese patients may present challenge for ultrasound guidance due to its poor visualization of deep anatomical structures. Regarding transforaminal epidural steroid injections, there are limited evidence to support that ultrasound guidance technology has equivalent efficacy and less complications comparing to fluoroscopy. However, further studies are required to prove the efficacy of ultrasound-guided transforaminal epidural injections. SI joint is unique due to its multiplanar orientation, irregular joint gap, partial ankylosis, and thick dorsal and interosseous ligament. Therefore, it can be difficult to access the joint space with fluoroscopic guidance and ultrasound guidance. CT scan, with its cross-sectional images, can identify posterior joint gap, is most likely the best guidance technology for this intervention. Intercostal nerves lie in the subcostal grove close to the plural space. Significant risk of pneumothorax is associated with intercostal blocks. Ultrasound can provide visualization of ribs and pleura. Therefore, it may improve the accuracy of the injection and reduce the risk of pneumothorax. At present time, most pain specialists are familiar with fluoroscopic guidance techniques, and fluoroscopic machines are readily available in the pain clinics. In the contrast, CT guidance can only be performed in specially equipped facilities. Ultrasound machine is generally portable and inexpensive in comparison to CT scanner and fluoroscopic machine. As pain specialists continue to improve their patient care, ultrasound and CT guidance will undoubtedly be incorporated more into the pain management practice. This review is based on a paucity of clinical evidence to compare these guidance technologies; clearly, more clinical studies is needed to further elucidate the pro and cons of each guidance method for various pain management interventions.

  2. Post-procedure bleeding in interventional radiology.

    PubMed

    Mayer, J; Tacher, V; Novelli, L; Djabbari, M; You, K; Chiaradia, M; Deux, J-F; Kobeiter, H

    2015-01-01

    Following interventional radiology procedures, bleeding can occur in 0.5 to 4% of the cases. Risk factors are related to the patient, to the procedure, and to the end organ. Bleeding is treated usually by interventional radiologists and consists mainly of embolization. Bleeding complications are preventable: before the procedure by checking hemostasis, during the procedure by ensuring the accurate puncture site (with ultrasound or fluoroscopy guidance) or by treating the puncture path using gelatin sponge, curaspon(®), biological glue or thermocoagulation, and after the procedure by carefully monitoring the patients. Copyright © 2015 Éditions françaises de radiologie. Published by Elsevier Masson SAS. All rights reserved.

  3. Technical advances of interventional fluoroscopy and flat panel image receptor.

    PubMed

    Lin, Pei-Jan Paul

    2008-11-01

    In the past decade, various radiation reducing devices and control circuits have been implemented on fluoroscopic imaging equipment. Because of the potential for lengthy fluoroscopic procedures in interventional cardiovascular angiography, these devices and control circuits have been developed for the cardiac catheterization laboratories and interventional angiography suites. Additionally, fluoroscopic systems equipped with image intensifiers have benefited from technological advances in x-ray tube, x-ray generator, and spectral shaping filter technologies. The high heat capacity x-ray tube, the medium frequency inverter generator with high performance switching capability, and the patient dose reduction spectral shaping filter had already been implemented on the image intensified fluoroscopy systems. These three underlying technologies together with the automatic dose rate and image quality (ADRIQ) control logic allow patients undergoing cardiovascular angiography procedures to benefit from "lower patient dose" with "high image quality." While photoconductor (or phosphor plate) x-ray detectors and signal capture thin film transistor (TFT) and charge coupled device (CCD) arrays are analog in nature, the advent of the flat panel image receptor allowed for fluoroscopy procedures to become more streamlined. With the analog-to-digital converter built into the data lines, the flat panel image receptor appears to become a digital device. While the transition from image intensified fluoroscopy systems to flat panel image receptor fluoroscopy systems is part of the on-going "digitization of imaging," the value of a flat panel image receptor may have to be evaluated with respect to patient dose, image quality, and clinical application capabilities. The advantage of flat panel image receptors has yet to be fully explored. For instance, the flat panel image receptor has its disadvantages as compared to the image intensifiers; the cost of the equipment is probably the most obvious. On the other hand, due to its wide dynamic range and linearity, lowering of patient dose beyond current practice could be achieved through the calibration process of the flat panel input dose rate being set to, for example, one half or less of current values. In this article various radiation saving devices and control circuits are briefly described. This includes various types of fluoroscopic systems designed to strive for reduction of patient exposure with the application of spectral shaping filters. The main thrust is to understand the ADRIQ control logic, through equipment testing, as it relates to clinical applications, and to show how this ADRIQ control logic "ties" those three technological advancements together to provide low radiation dose to the patient with high quality fluoroscopic images. Finally, rotational angiography with computed tomography (CT) and three dimensional (3-D) images utilizing flat panel technology will be reviewed as they pertain to diagnostic imaging in cardiovascular disease.

  4. Comparison of transradial and transfemoral artery approach for percutaneous coronary angiography and angioplasty: a retrospective seven-year experience from a north Indian center.

    PubMed

    Tewari, Satyendra; Sharma, Naveen; Kapoor, Aditya; Syal, Sanjeev Kumar; Kumar, Sudeep; Garg, Naveen; Goel, Pravin K

    2013-01-01

    With the increasing prevalence of coronary artery disease, percutaneous coronary artery procedures have become even more important. Our study has compared transradial to transfemoral artery approach for coronary procedures in Indian population. Comparison of transradial and transfemoral artery approach for percutaneous coronary procedures. 26,238 patients, who underwent percutaneous coronary artery procedures, were divided into two groups depending upon transradial and transfemoral artery approach and compared for the various demographic and clinical characteristics, risk factors profile, vascular access and procedural details. 26,238 patients underwent percutaneous coronary procedures at our center. 81% were male and 19% were female. 55.65% and 44.35% procedures were done through transfemoral and transradial approach, respectively. 17,417 (66.38%) coronary angiographies were done, out of which 53.92% were transradial and 46.08% were transfemoral procedures. 8821 (33.62%) Percutaneous Transluminal Coronary Angioplasty (PTCA) were done, out of which 25.46% and 74.54% were done through transradial and transfemoral approach, respectively. Mean fluoroscopy time was 4.40 ± 3.55 min for transradial and 3.30 ± 3.66 min for transfemoral CAG (p < 0.001). For PTCA mean fluoroscopy time was 13.53 ± 2.53 min for transradial and 12.61 ± 9.524 min for transfemoral PTCA (p < 0.001). Minor and major procedure related complications and total duration of hospital stay were lower in transradial as compared to transfemoral group. The number of percutaneous transradial procedures have increased significantly with reduced complication rates and comparable success rate to transfemoral approach, along with the additional benefits to patient in terms of patient comfort, preference and reduced cost of health delivery. Copyright © 2013 Cardiological Society of India. Published by Elsevier B.V. All rights reserved.

  5. Biplane interventional pediatric system with cone‐beam CT: dose and image quality characterization for the default protocols

    PubMed Central

    Vañó, Eliseo; Alejo, Luis; Ubeda, Carlos; Gutiérrez‐Larraya, Federico; Garayoa, Julia

    2016-01-01

    The aim of this study was to assess image quality and radiation dose of a biplane angiographic system with cone‐beam CT (CBCT) capability tuned for pediatric cardiac procedures. The results of this study can be used to explore dose reduction techniques. For pulsed fluoroscopy and cine modes, polymethyl methacrylate phantoms of various thicknesses and a Leeds TOR 18‐FG test object were employed. Various fields of view (FOV) were selected. For CBCT, the study employed head and body dose phantoms, Catphan 504, and an anthropomorphic cardiology phantom. The study also compared two 3D rotational angiography protocols. The entrance surface air kerma per frame increases by a factor of 3–12 when comparing cine and fluoroscopy frames. The biggest difference in the signal‐to‐noise ratio between fluoroscopy and cine modes occurs at FOV 32 cm because fluoroscopy is acquired at a 1440×1440 pixel matrix size and in unbinned mode, whereas cine is acquired at 720×720 pixels and in binned mode. The high‐contrast spatial resolution of cine is better than that of fluoroscopy, except for FOV 32 cm, because fluoroscopy mode with 32 cm FOV is unbinned. Acquiring CBCT series with a 16 cm head phantom using the standard dose protocol results in a threefold dose increase compared with the low‐dose protocol. Although the amount of noise present in the images acquired with the low‐dose protocol is much higher than that obtained with the standard mode, the images present better spatial resolution. A 1 mm diameter rod with 250 Hounsfield units can be distinguished in reconstructed images with an 8 mm slice width. Pediatric‐specific protocols provide lower doses while maintaining sufficient image quality. The system offers a novel 3D imaging mode. The acquisition of CBCT images results in increased doses administered to the patients, but also provides further diagnostic information contained in the volumetric images. The assessed CBCT protocols provide images that are noisy, but with very good spatial resolution. PACS number(s): 87.59.‐e, 87.59.‐C, 87.59.‐cf, 87.59.Dj, 87.57. uq PMID:27455474

  6. SU-E-I-26: Estimation of Micro-Angiographic Fluoroscope (MAF) Gain Settings for Digital Subtraction Angiography (DSA) Based on the Fluoroscopic Exposure.

    PubMed

    Ionita, C; Loughran, B; Nagesh, S Setlur; Jain, A; Bednarek, D; Rudin, S

    2012-06-01

    The MAF is a new high-resolution detector which is being clinically evaluated in neuro-vascular procedures. The detector contains a large-dynamic-range, high-sensitivity light image intensifier with variable gain. Since the MAF is a research prototype only partially integrated with the clinical system, x-ray technique parameters must be set manually. To improve workflow we developed an automatic method to estimate and set the proper LII voltage (MAF gain) for DSA acquisition based on the fluoroscopic parameters. The detector entrance exposure (XD) can be written as the x-ray tube output exposure (Xo) times an object attenuation factor and an inverse-square correction. If the object attenuation, scatter and distances are unchanged and the effect of x-ray kVp changes are neglected, then the DSA XD can be expressed as the ratio of Xo(DSA)/Xo(Fluoroscopy) multiplied with XD(fluoroscopy). We measured Xo for fluoroscopy and DSA for mAs and kVp ranges appropriate to neuro- vascular interventions and fit the data with a 2D function. To estimate the XD(Fluoroscopy) we derived a curve of XD versus LII-voltage for a mid- dynamic-range average pixel gray-level. Since the MAF system during clinical fluoroscopy automatically adjusts the LII voltage until the desired gray-level value is achieved, by reading that voltage we can estimate the XD(Fluoroscopy). Using the 2D-fit function, Xo(DSA) is automatically calculated for the kVp and mA values set and XD(DSA) can be estimated using the relation above. Using the inverse LII calibration curve, the proper LII-voltage can be determined for the desired average gray-level. The algorithm was implemented and evaluated in thirty-two in-vivo DSA runs on rabbits. The proper LII voltage was selected in all cases with no failures. Using the fluoroscopic LII gain setting to determine the appropriate DSA setting can greatly improve the workflow in clinical evaluations of the MAF. NIH Grants R01-EB008425, R01-EB002873 and an equipment grant from Toshiba Medical Systems Corp. © 2012 American Association of Physicists in Medicine.

  7. Biplane interventional pediatric system with cone-beam CT: dose and image quality characterization for the default protocols.

    PubMed

    Corredoira, Eva; Vañó, Eliseo; Alejo, Luis; Ubeda, Carlos; Gutiérrez-Larraya, Federico; Garayoa, Julia

    2016-07-08

    The aim of this study was to assess image quality and radiation dose of a biplane angiographic system with cone-beam CT (CBCT) capability tuned for pediatric cardiac procedures. The results of this study can be used to explore dose reduction techniques. For pulsed fluoroscopy and cine modes, polymethyl methacrylate phantoms of various thicknesses and a Leeds TOR 18-FG test object were employed. Various fields of view (FOV) were selected. For CBCT, the study employed head and body dose phantoms, Catphan 504, and an anthropomorphic cardiology phantom. The study also compared two 3D rotational angiography protocols. The entrance surface air kerma per frame increases by a factor of 3-12 when comparing cine and fluoroscopy frames. The biggest difference in the signal-to- noise ratio between fluoroscopy and cine modes occurs at FOV 32 cm because fluoroscopy is acquired at a 1440 × 1440 pixel matrix size and in unbinned mode, whereas cine is acquired at 720 × 720 pixels and in binned mode. The high-contrast spatial resolution of cine is better than that of fluoroscopy, except for FOV 32 cm, because fluoroscopy mode with 32 cm FOV is unbinned. Acquiring CBCT series with a 16 cm head phantom using the standard dose protocol results in a threefold dose increase compared with the low-dose protocol. Although the amount of noise present in the images acquired with the low-dose protocol is much higher than that obtained with the standard mode, the images present better spatial resolution. A 1 mm diameter rod with 250 Hounsfield units can be distinguished in reconstructed images with an 8 mm slice width. Pediatric-specific protocols provide lower doses while maintaining sufficient image quality. The system offers a novel 3D imaging mode. The acquisition of CBCT images results in increased doses administered to the patients, but also provides further diagnostic information contained in the volumetric images. The assessed CBCT protocols provide images that are noisy, but with very good spatial resolution. © 2016 The Authors.

  8. An experimental animal model for percutaneous procedures used in trigeminal neuralgia.

    PubMed

    Herta, Johannes; Wang, Wei-Te; Höftberger, Romana; Breit, Sabine; Kneissl, Sibylle; Bergmeister, Helga; Ferraz-Leite, Heber

    2017-07-01

    This study describes an experimental rabbit model that allows the reproduction of percutaneous operations that are used in patients with trigeminal neuralgia (TN). Attention was given to an exact anatomical description of the rabbit's middle cranial fossa as well as the establishment of conditions for a successful procedure. Morphometric measurements were taken from 20 rabbit skulls and CT scans. The anatomy of the trigeminal nerve, as well as its surrounding structures, was assessed by bilateral dissection of 13 New Zealand white rabbits (NWR). An ideal approach of placing a needle through the foramen ovale to reach the TG was sought. Validation of correct placement was realized by fluoroscopy and confirmed by dissection. Precise instructions for successful reproduction of percutaneous procedures in NWR were described. According to morphological measurements, for balloon compression of the trigeminal ganglion (TG) the maximal diameter of an introducing cannula is 1.85 mm. The diameter of an empty balloon catheter should not exceed 1.19 mm, and the length of the inflatable part of the balloon can range up to 4 mm. For thermocoagulation the needle electrodes must not exceed an external diameter of 1.39, mm and the length of the non-insolated tip can range up to 4 mm. Glycerol rhizolysis can be achieved because the trigeminal cistern in the NWR is a closed space that allows a long dwelling time (>10 min) of the contrast agent. An experimental NWR model intended for the reproduction of percutaneous procedures on the TG has been meticulously described. This provides a tool that enables further standardized animal research in the field of surgical treatment of TN.

  9. Use of asymmetric bidirectional catheters with different curvature radius for catheter ablation of cardiac arrhythmias.

    PubMed

    Mantziari, Lilian; Suman-Horduna, Irina; Gujic, Marko; Jones, David G; Wong, Tom; Markides, Vias; Foran, John P; Ernst, Sabine

    2013-06-01

    The impact of recently introduced asymmetric bidirectional ablation catheters on procedural parameters and acute success rates of ablation procedures is unknown. We retrospectively analyzed data regarding ablations using a novel bidirectional catheter in a tertiary cardiac center and compared these in 1:5 ratio with a control group of procedures matched for age, gender, operator, and ablation type. A total of 50 cases and 250 controls of median age 60 (50-68) years were studied. Structural heart disease was equally prevalent in both groups (39%) while history of previous ablations was more common in the study arm (54% vs 30%, P = 0.001). Most of the ablation cases were for atrial fibrillation (46%), followed by atrial tachycardia (28%), supraventricular tachycardia (12%), and ventricular tachycardia (14%). Median procedure duration was 128 (52-147) minutes with the bidirectional, versus 143 (105-200) minutes with the conventional catheter (P = 0.232), and median fluoroscopy time was 17 (10-34) minutes versus 23 (12-39) minutes, respectively (P = 0.988). There was a trend toward a lower procedure duration for the atrial tachycardia ablations, 89 (52-147) minutes versus 130 (100-210) minutes, P = 0.064. The procedure was successfully completed in 96% of the bidirectional versus 84% of the control cases (P = 0.151). A negative correlation was observed between the relative fluoroscopy duration and the case number (r = -0.312, P = 0.028), reflecting the learning curve for the bidirectional catheter. The introduction of the bidirectional catheter resulted in no prolongation of procedure parameters and similar success rates, while there was a trend toward a lower procedure duration for atrial tachycardia ablations. ©2013, The Authors. Journal compilation ©2013 Wiley Periodicals, Inc.

  10. Minilaparotomy-Assisted Transmesenteric-Transjugular Intrahepatic Portosystemic Shunt: Comparison with Conventional Transjugular Approach

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jalaeian, Hamed, E-mail: hjalaeia@umn.edu; Talaie, Reza; D’Souza, Donna

    PurposeThis study was performed to compare the intrahepatic shunt function outcome and procedural complications of minilaparotomy-assisted transmesenteric (MAT)-transjugular intrahepatic portosystemic shunt (TIPS) placement with the conventional transjugular method.MethodsThis is a retrospective review of all patients who had a MAT or conventional TIPS procedure over a 6-year period at our institute. The primary patency rate, fluoroscopy time, technical success, major procedure-related complications, and mortality data were compared between two treatment groups.ResultsWe included 49 patients with MAT-TIPS, and 63 with conventional TIPS, with an average follow-up of 21.43 months. The primary patency rates at 6 and 12 months were 82.9 and 66.7 % in themore » conventional TIPS group, and 81.0 and 76.5 % in the MAT-TIPS group (p = 1.000, and 0.529), respectively. There was no significant difference in technical success rate, post-procedure portosystemic pressure gradient, fluoroscopy time, and peri-procedural mortality rate between treatment groups. Major procedural-related complications were seen more frequently among MAT-TIPS patients (p = 0.012). In the MAT-TIPS group, 5 (10.2 %) patients developed post-procedure minilaparotomy, wound-related complications, and 5 (10.2 %) developed bacterial peritonitis; whereas, none of patients with conventional TIPS had either of these complications (p = 0.014).ConclusionWhile both MAT-TIPS and conventional TIPS had similar shunt primary patency rate and technical success rate, the MAT approach was associated with a significantly higher rate of minilaparotomy-related wound complications or infectious complications. These complications maybe prevented by a change in post-procedure monitoring and therapy.« less

  11. Single Center Retrospective Analysis of Conventional and Radial TIG Catheters for Transradial Diagnostic Coronary Angiography.

    PubMed

    Vorpahl, Marc; Koehler, Till; Foerst, Jason; Panagiotopoulos, Spyridon; Schleiting, Heinrich; Koss, Klaus; Ziegler, Gunda; Brinkmann, Hilmar; Seyfarth, Melchior; Tiroch, Klaus

    2015-01-01

    Current guidelines favor the radial approach for coronary angiography. Therefore, specialty radial diagnostic catheters were designed to engage both coronary arteries with a single device. However, it is unclear if single catheters are superior to conventional catheters. A retrospective analysis was performed of consecutive right radial coronary angiographies to determine catheter use, fluoroscopy time, radiation dosage, and consumption of contrast. Procedures were performed with a single TIG catheter or conventional catheters (CONV). Procedures with coronary artery bypass grafts or ventricular angiographies were excluded. 273 transradial procedures were performed successfully. 95 procedures were performed with CONV and 178 procedures with a TIG. Crossover to additional catheters was higher in TIG (15.2%) compared to CONV (5.3%, p = 0.02). Fluoroscopy time was comparable between CONV and TIG, without crossover (2.2 ± 1.2 min versus 2.3 ± 1.2 min; n.s.), however, greater in the case of crossover for CONV (5.8 ± 0.7) and TIG (7.6 ± 3.0; p = 0.0001). Radiation dosage was similar in CONV and the TIG, without crossover (1419 ± 1075, cGy∗cm(2) versus 1690 ± 1138; n.s.), however, greater for CONV (2374 ± 620) and TIG (3733 ± 2281, p = 0.05) with crossover. Overall, the amount of contrast was greater in TIG (56 ± 13 mL) versus CONV (48 ± 3 mL; p = 0.0003). CONV femoral catheters may be the primary choice for radial approach.

  12. Radiation exposure to the eye lens of orthopaedic surgeons during various orthopaedic procedures.

    PubMed

    Romanova, K; Vassileva, J; Alyakov, M

    2015-07-01

    The aim of the present study was to assess the radiation dose to the eye lens of orthopaedic surgeons during various orthopaedic procedures and to make efforts to ensure that radiation protection is optimised. The study was performed for Fractura femoris and Fractura cruris procedures performed in orthopaedic operating theatres, as well as for fractures of wrist, ankle and hand/shoulder performed in the emergency trauma room. The highest mean value of the eye lens dose of 47.2 μSv and higher mean fluoroscopy time of 3 min, as well as the corresponding highest maximum values of 77.1 μSv and 5.0 min were observed for the Fractura femoris procedure performed with the Biplanar 500e fluoroscopy systems. At a normal workload, the estimated mean annual dose values do not exceed the annual occupational dose limit for the lens of eye, but at a heavy workload in the department, this dose limit could be achieved or exceeded. The use of protective lead glasses is recommended as they could reduce the radiation exposure of the lens of the eye. The phantom measurements demonstrated that the use of half-dose mode could additionally reduce dose to the operator's eye lens. © The Author 2015. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  13. Radiation exposure and safety practices during pediatric central line placement

    PubMed Central

    Saeman, Melody R.; Burkhalter, Lorrie S.; Blackburn, Timothy J.; Murphy, Joseph T.

    2015-01-01

    Purpose Pediatric surgeons routinely use fluoroscopy for central venous line (CVL) placement. We examined radiation safety practices and patient/surgeon exposure during fluoroscopic CVL. Methods Fluoroscopic CVL procedures performed by 11 pediatric surgeons in 2012 were reviewed. Fluoroscopic time (FT), patient exposure (mGy), and procedural data were collected. Anthropomorphic phantom simulations were used to calculate scatter and dose (mSv). Surgeons were surveyed regarding safety practices. Results 386 procedures were reviewed. Median FT was 12.8 seconds. Median patient estimated effective dose was 0.13 mSv. Median annual FT per surgeon was 15.4 minutes. Simulations showed no significant difference (p = 0.14) between reported exposures (median 3.5 mGy/min) and the modeled regression exposures from the C-arm default mode (median 3.4 mGy/min). Median calculated surgeon exposure was 1.5 mGy/year. Eight of 11 surgeons responded to the survey. Only three reported 100% lead protection and frequent dosimeter use. Conclusion We found non-standard radiation training, safety practices, and dose monitoring for the 11 surgeons. Based on simulations, the C-arm default setting was typically used instead of low dose. While most CVL procedures have low patient/surgeon doses, every effort should be used to minimize patient and occupational exposure, suggesting the need for formal hands-on training for non-radiologist providers using fluoroscopy. PMID:25837269

  14. Prospective comparison of virtual fluoroscopy to fluoroscopy and plain radiographs for placement of lumbar pedicle screws.

    PubMed

    Resnick, Daniel K

    2003-06-01

    Fluoroscopy-based frameless stereotactic systems provide feedback to the surgeon using virtual fluoroscopic images. The real-life accuracy of these virtual images has not been compared with traditional fluoroscopy in a clinical setting. We prospectively studied 23 consecutive cases. In two cases, registration errors precluded the use of virtual fluoroscopy. Pedicle probes placed with virtual fluoroscopic imaging were imaged with traditional fluoroscopy in the remaining 21 cases. Position of the probes was judged to be ideal, acceptable but not ideal, or not acceptable based on the traditional fluoroscopic images. Virtual fluoroscopy was used to place probes in for 97 pedicles from L1 to the sacrum. Eighty-eight probes were judged to be in ideal position, eight were judged to be acceptable but not ideal, and one probe was judged to be in an unacceptable position. This probe was angled toward an adjacent disc space. Therefore, 96 of 97 probes placed using virtual fluoroscopy were found to be in an acceptable position. The positive predictive value for acceptable screw placement with virtual fluoroscopy compared with traditional fluoroscopy was 99%. A probe placed with virtual fluoroscopic guidance will be judged to be in an acceptable position when imaged with traditional fluoroscopy 99% of the time.

  15. Exposure of medical staff to radiation during osteosynthesis of proximal femoral fracture: descriptive analysis and comparison of different devices.

    PubMed

    Siedlecki, Cédric; Gauthé, Rémi; Gillibert, André; Bellenger, Kevin; Roussignol, Xavier; Ould-Slimane, Mourad

    2017-10-01

    The use of fluoroscopy is necessary during proximal femoral fracture (PFF) osteosynthesis. The frequency of these procedures justifies a description of radiation exposure and comparisons between different techniques and between the different surgical team members. This observational prospective and comparative study includes a series of 68 patients with PFF receiving osteosynthesis. Radiation exposure was assessed for all members of the operating team. The radiation dose measurements for the different members of the surgical team during PFF osteosynthesis were compared. The factors affecting the radiation dose were investigated. The mean active dosimeter readings for each operation were 7.39 µSv for the primary surgeon, 3.93 µSv for the assistant surgeon, 1.92 µSv for the instrument nurse, 1.25 µSv for the circulating nurse, and 0.64 µSv for the anaesthesiologist, respectively. Doses decreased significantly between these different members of the medical team (all p < 0.001). The dose also varied with patient age and BMI, as well as with fluoroscopy time and operating time, but not with type of fracture or type of osteosynthesis. Medical staff receives significantly different doses depending on their position in relation to the radiation source. Operating time and fluoroscopy time are the modifiable factors that affect the radiation dose. The radiation doses received by the different members of the medical teams involved in proximal femur osteosynthesis procedures all fall below the doses recommended by the International Commission on Radiation Units and Measurements.

  16. Modified fluoroscopy-guided sacroiliac joint injection: a technical report.

    PubMed

    Liliang, Po-Chou; Liang, Cheng-Loong; Lu, Kang; Weng, Hui-Ching; Syu, Fei-Kai

    2014-09-01

    Sacroiliac joint (SIJ) injection can occasionally be challenging. We describe our experience in using conventional technique, and we developed an adjustment to overcome difficulties incurred. Conventional technique required superimposition of the posterior and anterior SIJ lines. If this technique failed to provide entry into the joint, fluoroscopy was slightly adjusted to obtain an oblique view. Of 50 SIJ injections, 29 (58%; 44-72%) were successfully performed using conventional technique. In another 21 procedures, 18 (85.7%; 64-99%) were subsequently completed using oblique view technique. The medial joint line, viewed from this angle, corresponded to the posterior joint line in 17 cases. The lateral joint line corresponded to the posterior joint line in one case. Oblique view technique can improve the success rate of SIJ injection. Wiley Periodicals, Inc.

  17. Training safer orthopedic surgeons. Construct validation of a virtual-reality simulator for hip fracture surgery.

    PubMed

    Akhtar, Kashif; Sugand, Kapil; Sperrin, Matthew; Cobb, Justin; Standfield, Nigel; Gupte, Chinmay

    2015-01-01

    Virtual-reality (VR) simulation in orthopedic training is still in its infancy, and much of the work has been focused on arthroscopy. We evaluated the construct validity of a new VR trauma simulator for performing dynamic hip screw (DHS) fixation of a trochanteric femoral fracture. 30 volunteers were divided into 3 groups according to the number of postgraduate (PG) years and the amount of clinical experience: novice (1-4 PG years; less than 10 DHS procedures); intermediate (5-12 PG years; 10-100 procedures); expert (> 12 PG years; > 100 procedures). Each participant performed a DHS procedure and objective performance metrics were recorded. These data were analyzed with each performance metric taken as the dependent variable in 3 regression models. There were statistically significant differences in performance between groups for (1) number of attempts at guide-wire insertion, (2) total fluoroscopy time, (3) tip-apex distance, (4) probability of screw cutout, and (5) overall simulator score. The intermediate group performed the procedure most quickly, with the lowest fluoroscopy time, the lowest tip-apex distance, the lowest probability of cutout, and the highest simulator score, which correlated with their frequency of exposure to running the trauma lists for hip fracture surgery. This study demonstrates the construct validity of a haptic VR trauma simulator with surgeons undertaking the procedure most frequently performing best on the simulator. VR simulation may be a means of addressing restrictions on working hours and allows trainees to practice technical tasks without putting patients at risk. The VR DHS simulator evaluated in this study may provide valid assessment of technical skill.

  18. Interventional robotic systems: Applications and technology state-of-the-art

    PubMed Central

    CLEARY, KEVIN; MELZER, ANDREAS; WATSON, VANCE; KRONREIF, GERNOT; STOIANOVICI, DAN

    2011-01-01

    Many different robotic systems have been developed for invasive medical procedures. In this article we will focus on robotic systems for image-guided interventions such as biopsy of suspicious lesions, interstitial tumor treatment, or needle placement for spinal blocks and neurolysis. Medical robotics is a young and evolving field and the ultimate role of these systems has yet to be determined. This paper presents four interventional robotics systems designed to work with MRI, CT, fluoroscopy, and ultrasound imaging devices. The details of each system are given along with any phantom, animal, or human trials. The systems include the AcuBot for active needle insertion under CT or fluoroscopy, the B-Rob systems for needle placement using CT or ultrasound, the INNOMOTION for MRI and CT interventions, and the MRBot for MRI procedures. Following these descriptions, the technology issues of image compatibility, registration, patient movement and respiration, force feedback, and control mode are briefly discussed. It is our belief that robotic systems will be an important part of future interventions, but more research and clinical trials are needed. The possibility of performing new clinical procedures that the human cannot achieve remains an ultimate goal for medical robotics. Engineers and physicians should work together to create and validate these systems for the benefits of patients everywhere. PMID:16754193

  19. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Matsui, Yusuke, E-mail: wckyh140@yahoo.co.jp; Hiraki, Takao, E-mail: takaoh@tc4.so-net.ne.jp; Gobara, Hideo, E-mail: gobara@cc.okayama-u.ac.jp

    IntroductionComputed tomography (CT) fluoroscopy-guided renal cryoablation and lung radiofrequency ablation (RFA) have received increasing attention as promising cancer therapies. Although radiation exposure of interventional radiologists during these procedures is an important concern, data on operator exposure are lacking.Materials and MethodsRadiation dose to interventional radiologists during CT fluoroscopy-guided renal cryoablation (n = 20) and lung RFA (n = 20) was measured prospectively in a clinical setting. Effective dose to the operator was calculated from the 1-cm dose equivalent measured on the neck outside the lead apron, and on the left chest inside the lead apron, using electronic dosimeters. Equivalent dose to the operator’s finger skinmore » was measured using thermoluminescent dosimeter rings.ResultsThe mean (median) effective dose to the operator per procedure was 6.05 (4.52) μSv during renal cryoablation and 0.74 (0.55) μSv during lung RFA. The mean (median) equivalent dose to the operator’s finger skin per procedure was 2.1 (2.1) mSv during renal cryoablation, and 0.3 (0.3) mSv during lung RFA.ConclusionRadiation dose to interventional radiologists during renal cryoablation and lung RFA were at an acceptable level, and in line with recommended dose limits for occupational radiation exposure.« less

  20. Maximum voltage gradient technique for optimization of ablation for typical atrial flutter with zero-fluoroscopy approach

    PubMed Central

    Deutsch, Karol; Śledź, Janusz; Mazij, Mariusz; Ludwik, Bartosz; Labus, Michał; Karbarz, Dariusz; Pasicka, Bernadetta; Chrabąszcz, Michał; Śledź, Arkadiusz; Klank-Szafran, Monika; Vitali-Sendoz, Laura; Kameczura, Tomasz; Śpikowski, Jerzy; Stec, Piotr; Ujda, Marek; Stec, Sebastian

    2017-01-01

    Abstract Radiofrequency catheter ablation (RFCA) is an established effective method for the treatment of typical cavo-tricuspid isthmus (CTI)-dependent atrial flutter (AFL). The introduction of 3-dimensional electro-anatomic systems enables RFCA without fluoroscopy (No-X-Ray [NXR]). The aim of this study was to evaluate the feasibility and effectiveness of CTI RFCA during implementation of the NXR approach and the maximum voltage-guided (MVG) technique for ablation of AFL. Data were obtained from prospective standardized multicenter ablation registry. Consecutive patients with the first RFCA for CTI-dependent AFL were recruited. Two navigation approaches (NXR and fluoroscopy based as low as reasonable achievable [ALARA]) and 2 mapping and ablation techniques (MVG and pull-back technique [PBT]) were assessed. NXR + MVG (n  =  164; age: 63.7 ± 9.5; 30% women), NXR + PBT (n  =  55; age: 63.9 ± 10.7; 39% women); ALARA + MVG (n  =  36; age: 64.2 ± 9.6; 39% women); and ALARA + PBT (n  =  205; age: 64.7 ± 9.1; 30% women) were compared, respectively. All groups were simplified with a 2-catheter femoral approach using 8-mm gold tip catheters (Osypka AG, Germany or Biotronik, Germany) with 15 min of observation. The MVG technique was performed using step-by-step application by mapping the largest atrial signals within the CTI. Bidirectional block in CTI was achieved in 99% of all patients (P  =  NS, between groups). In NXR + MVG and NXR + PBT groups, the procedure time decreased (45.4 ± 17.6 and 47.2 ± 15.7 min vs. 52.6 ± 23.7 and 59.8 ± 24.0 min, P < .01) as compared to ALARA + MVG and ALARA + PBT subgroups. In NXR + MVG and NXR + PBT groups, 91% and 98% of the procedures were performed with complete elimination of fluoroscopy. The NXR approach was associated with a significant reduction in fluoroscopy exposure (from 0.2 ± 1.1 [NXR + PBT] and 0.3 ± 1.6 [NXR + MVG] to 7.7 ± 6.0 min [ALARA + MVG] and 9.1 ± 7.2 min [ALARA + PBT], P < .001). The total application time significantly decreased in the MVG technique subgroup both in NXR and ALARA (P < .01). No major complications were observed in either groups. Complete elimination of fluoroscopy is feasible, safe, and effective during RFCA of CTI in almost all AFL patients without cardiac implanted electronic devices. The most optimal method for RFCA of CTI-dependent AFL seems to be MVG; however, it required validation of optimal RFCA's parameters with clinical follow-up. PMID:28640075

  1. Maximum voltage gradient technique for optimization of ablation for typical atrial flutter with zero-fluoroscopy approach.

    PubMed

    Deutsch, Karol; Śledź, Janusz; Mazij, Mariusz; Ludwik, Bartosz; Labus, Michał; Karbarz, Dariusz; Pasicka, Bernadetta; Chrabąszcz, Michał; Śledź, Arkadiusz; Klank-Szafran, Monika; Vitali-Sendoz, Laura; Kameczura, Tomasz; Śpikowski, Jerzy; Stec, Piotr; Ujda, Marek; Stec, Sebastian

    2017-06-01

    Radiofrequency catheter ablation (RFCA) is an established effective method for the treatment of typical cavo-tricuspid isthmus (CTI)-dependent atrial flutter (AFL). The introduction of 3-dimensional electro-anatomic systems enables RFCA without fluoroscopy (No-X-Ray [NXR]). The aim of this study was to evaluate the feasibility and effectiveness of CTI RFCA during implementation of the NXR approach and the maximum voltage-guided (MVG) technique for ablation of AFL.Data were obtained from prospective standardized multicenter ablation registry. Consecutive patients with the first RFCA for CTI-dependent AFL were recruited. Two navigation approaches (NXR and fluoroscopy based as low as reasonable achievable [ALARA]) and 2 mapping and ablation techniques (MVG and pull-back technique [PBT]) were assessed. NXR + MVG (n  =  164; age: 63.7 ± 9.5; 30% women), NXR + PBT (n  =  55; age: 63.9 ± 10.7; 39% women); ALARA + MVG (n  =  36; age: 64.2 ± 9.6; 39% women); and ALARA + PBT (n  =  205; age: 64.7 ± 9.1; 30% women) were compared, respectively. All groups were simplified with a 2-catheter femoral approach using 8-mm gold tip catheters (Osypka AG, Germany or Biotronik, Germany) with 15 min of observation. The MVG technique was performed using step-by-step application by mapping the largest atrial signals within the CTI.Bidirectional block in CTI was achieved in 99% of all patients (P  =  NS, between groups). In NXR + MVG and NXR + PBT groups, the procedure time decreased (45.4 ± 17.6 and 47.2 ± 15.7 min vs. 52.6 ± 23.7 and 59.8 ± 24.0 min, P < .01) as compared to ALARA + MVG and ALARA + PBT subgroups. In NXR + MVG and NXR + PBT groups, 91% and 98% of the procedures were performed with complete elimination of fluoroscopy. The NXR approach was associated with a significant reduction in fluoroscopy exposure (from 0.2 ± 1.1 [NXR + PBT] and 0.3 ± 1.6 [NXR + MVG] to 7.7 ± 6.0 min [ALARA + MVG] and 9.1 ± 7.2 min [ALARA + PBT], P < .001). The total application time significantly decreased in the MVG technique subgroup both in NXR and ALARA (P < .01). No major complications were observed in either groups.Complete elimination of fluoroscopy is feasible, safe, and effective during RFCA of CTI in almost all AFL patients without cardiac implanted electronic devices. The most optimal method for RFCA of CTI-dependent AFL seems to be MVG; however, it required validation of optimal RFCA's parameters with clinical follow-up.

  2. Fluoroscopy-Guided Pull-Through Gastrostomy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Pitton, M. B., E-mail: pitton@radiologie.klinik.uni-mainz.de; Herber, S.; Dueber, C.

    2008-01-15

    The purpose of this study was to simplify a fluoroscopy guided gastrostomy technique using pull-type tubes which are traditionally introduced with gastroscopic assistance. The stomach was transorally probed with a 5-Fr catheter and a guidewire. A second access was performed percutaneously through the anterior abdominal and gastric wall using an 8-Fr sheath and an 8-Fr guiding catheter. A duplicated guidewire was introduced through the guiding catheter in order to result in a great custom-made loop within the stomach. The transoral guidewire was captured and tightened with this loop and the guiding catheter, and both were subsequently pulled by the transoralmore » guidewire until the tip of the guiding catheter exited the mouth. A thread was fed through the guiding catheter for fixation of the pull-type gastrostomy tube. Finally, the fixed tube was pulled through the esophagus into the stomach and through the abdominal wall until the anterior gastric wall fixed the retention plate of the tube. Thirty-seven patients (28 male, 9 female; age, 65.1 {+-} 14.4 years) with miscellaneous indications for percutaneous gastrostomies were supplied with pull-type gastrostomy catheters in a fluoroscopy technique without endoscopic assistance. Twenty-five of the 37 patients (67.6%) had undergone unsuccessful preceding gastroscopically guided PEG attempts because of tumor stenosis (n = 12) or impossible transillumination of the abdominal wall (n = 13). All procedures were technically successful, without major complications. Particularly, all patients with frustrating gastroscopic attempts were successfully provided with pull-type gastrostomy tubes. Five minor complications occurred: one tube loss during the passage of the hypopoharynx because of a torn thread, one transient small leakage alongside the tube (both successfully treated), and three cases of transient moderate local pain without leakage (symptomatic treatment). We conclude that this fluoroscopy-guided pull-through gastrostomy technique is easy and safe to perform and may be suggested as a standard procedure for radiological gastrostomies. It combines the ease of the radiological approach with the advantages of the pull-type tube devices, particularly the benefits of the typical retention plates.« less

  3. Image-guided interventional procedures in the dog and cat.

    PubMed

    Vignoli, Massimo; Saunders, Jimmy H

    2011-03-01

    Medical imaging is essential for the diagnostic workup of many soft tissue and bone lesions in dogs and cats, but imaging modalities do not always allow the clinician to differentiate inflammatory or infectious conditions from neoplastic disorders. This review describes interventional procedures in dogs and cats for collection of samples for cytological or histopathological examinations under imaging guidance. It describes the indications and procedures for imaging-guided sampling, including ultrasound (US), computed tomography (CT), magnetic resonance imaging and fluoroscopy. US and CT are currently the modalities of choice in interventional imaging. Copyright © 2009 Elsevier Ltd. All rights reserved.

  4. Comparison of transradial and transfemoral artery approach for percutaneous coronary angiography and angioplasty: A retrospective seven-year experience from a north Indian center

    PubMed Central

    Tewari, Satyendra; Sharma, Naveen; Kapoor, Aditya; Syal, Sanjeev Kumar; Kumar, Sudeep; Garg, Naveen; Goel, Pravin K.

    2013-01-01

    Background With the increasing prevalence of coronary artery disease, percutaneous coronary artery procedures have become even more important. Our study has compared transradial to transfemoral artery approach for coronary procedures in Indian population. Aims and objective Comparison of transradial and transfemoral artery approach for percutaneous coronary procedures. Material & methods 26,238 patients, who underwent percutaneous coronary artery procedures, were divided into two groups depending upon transradial and transfemoral artery approach and compared for the various demographic and clinical characteristics, risk factors profile, vascular access and procedural details. Results 26,238 patients underwent percutaneous coronary procedures at our center. 81% were male and 19% were female. 55.65% and 44.35% procedures were done through transfemoral and transradial approach, respectively. 17,417 (66.38%) coronary angiographies were done, out of which 53.92% were transradial and 46.08% were transfemoral procedures. 8821 (33.62%) Percutaneous Transluminal Coronary Angioplasty (PTCA) were done, out of which 25.46% and 74.54% were done through transradial and transfemoral approach, respectively. Mean fluoroscopy time was 4.40 ± 3.55 min for transradial and 3.30 ± 3.66 min for transfemoral CAG (p < 0.001). For PTCA mean fluoroscopy time was 13.53 ± 2.53 min for transradial and 12.61 ± 9.524 min for transfemoral PTCA (p < 0.001). Minor and major procedure related complications and total duration of hospital stay were lower in transradial as compared to transfemoral group. Conclusion The number of percutaneous transradial procedures have increased significantly with reduced complication rates and comparable success rate to transfemoral approach, along with the additional benefits to patient in terms of patient comfort, preference and reduced cost of health delivery. PMID:23992998

  5. A novel drill design for photoacoustic guided surgeries

    NASA Astrophysics Data System (ADS)

    Shubert, Joshua; Lediju Bell, Muyinatu A.

    2018-02-01

    Fluoroscopy is currently the standard approach for image guidance of surgical drilling procedures. In addition to the harmful radiation dose to the patient and surgeon, fluoroscopy fails to visualize critical structures such as blood vessels and nerves within the drill path. Photoacoustic imaging is a well-suited imaging method to visualize these structures and it does not require harmful ionizing radiation. However, there is currently no clinical system available to deliver light to occluded drill bit tips. To address this challenge, a prototype drill was designed, built, and tested using an internal light delivery system that allows laser energy to be transferred from a stationary laser source to the tip of a spinning drill bit. Photoacoustic images were successfully obtained with the drill bit submerged in water and with the drill tip inserted into a thoracic vertebra from a human cadaver.

  6. Fast 3-T MR-guided transrectal prostate biopsy using an in-room tablet device for needle guide alignment: a feasibility study.

    PubMed

    Overduin, Christiaan G; Heidkamp, Jan; Rothgang, Eva; Barentsz, Jelle O; de Lange, Frank; Fütterer, Jurgen J

    2018-05-22

    To assess the feasibility of adding a tablet device inside the scanner room to assist needle-guide alignment during magnetic resonance (MR)-guided transrectal prostate biopsy. Twenty patients with one cancer-suspicious region (CSR) with PI-RADS score ≥ 4 on diagnostic multiparametric MRI were prospectively enrolled. Two orthogonal scan planes of an MR fluoroscopy sequence (~3 images/s) were aligned to the CSR and needle-guide pivoting point. Targeting was achieved by manipulating the needle-guide under MR fluoroscopy feedback on the in-room tablet device. Technical feasibility and targeting success were assessed. Complications and biopsy procedure times were also recorded. Needle-guide alignment with the in-room tablet device was technically successful in all patients and allowed sampling after a single alignment step in 19/20 (95%) CSRs (median size 14 mm, range: 4-45). Biopsy cores contained cancer in 18/20 patients. There were no per-procedural or post-biopsy complications. Using the tablet device, the mean time to first biopsy was 5.8 ± 1.0 min and the mean total procedure time was 23.7 ± 4.1 min. Use of an in-room tablet device to assist needle-guide alignment was feasible and safe during MR-guided transrectal prostate biopsy. Initial experience indicates potential for procedure time reduction. • Performing MR-guided prostate biopsy using an in-room tablet device is feasible. • CSRs could be sampled after a single alignment step in 19/20 patients. • The mean procedure time for biopsy with the tablet device was 23.7 min.

  7. Use of Intracardiac Echocardiography in Interventional Cardiology: Working With the Anatomy Rather Than Fighting It.

    PubMed

    Enriquez, Andres; Saenz, Luis C; Rosso, Raphael; Silvestry, Frank E; Callans, David; Marchlinski, Francis E; Garcia, Fermin

    2018-05-22

    The indications for catheter-based structural and electrophysiological procedures have recently expanded to more complex scenarios, in which an accurate definition of the variable individual cardiac anatomy is key to obtain optimal results. Intracardiac echocardiography (ICE) is a unique imaging modality able to provide high-resolution real-time visualization of cardiac structures, continuous monitoring of catheter location within the heart, and early recognition of procedural complications, such as pericardial effusion or thrombus formation. Additional benefits are excellent patient tolerance, reduction of fluoroscopy time, and lack of need for general anesthesia or a second operator. For these reasons, ICE has largely replaced transesophageal echocardiography as ideal imaging modality for guiding certain procedures, such as atrial septal defect closure and catheter ablation of cardiac arrhythmias, and has an emerging role in others, including mitral valvuloplasty, transcatheter aortic valve replacement, and left atrial appendage closure. In electrophysiology procedures, ICE allows integration of real-time images with electroanatomic maps; it has a role in assessment of arrhythmogenic substrate, and it is particularly useful for mapping structures that are not visualized by fluoroscopy, such as the interatrial or interventricular septum, papillary muscles, and intracavitary muscular ridges. Most recently, a three-dimensional (3D) volumetric ICE system has also been developed, with potential for greater anatomic information and a promising role in structural interventions. In this state-of-the-art review, we provide guidance on how to conduct a comprehensive ICE survey and summarize the main applications of ICE in a variety of structural and electrophysiology procedures. © 2018 American Heart Association, Inc.

  8. Single Center Retrospective Analysis of Conventional and Radial TIG Catheters for Transradial Diagnostic Coronary Angiography

    PubMed Central

    Vorpahl, Marc; Koehler, Till; Foerst, Jason; Panagiotopoulos, Spyridon; Schleiting, Heinrich; Koss, Klaus; Ziegler, Gunda; Brinkmann, Hilmar; Seyfarth, Melchior; Tiroch, Klaus

    2015-01-01

    Current guidelines favor the radial approach for coronary angiography. Therefore, specialty radial diagnostic catheters were designed to engage both coronary arteries with a single device. However, it is unclear if single catheters are superior to conventional catheters. A retrospective analysis was performed of consecutive right radial coronary angiographies to determine catheter use, fluoroscopy time, radiation dosage, and consumption of contrast. Procedures were performed with a single TIG catheter or conventional catheters (CONV). Procedures with coronary artery bypass grafts or ventricular angiographies were excluded. 273 transradial procedures were performed successfully. 95 procedures were performed with CONV and 178 procedures with a TIG. Crossover to additional catheters was higher in TIG (15.2%) compared to CONV (5.3%, p = 0.02). Fluoroscopy time was comparable between CONV and TIG, without crossover (2.2 ± 1.2 min versus 2.3 ± 1.2 min; n.s.), however, greater in the case of crossover for CONV (5.8 ± 0.7) and TIG (7.6 ± 3.0; p = 0.0001). Radiation dosage was similar in CONV and the TIG, without crossover (1419 ± 1075, cGy∗cm2 versus 1690 ± 1138; n.s.), however, greater for CONV (2374 ± 620) and TIG (3733 ± 2281, p = 0.05) with crossover. Overall, the amount of contrast was greater in TIG (56 ± 13 mL) versus CONV (48 ± 3 mL; p = 0.0003). CONV femoral catheters may be the primary choice for radial approach. PMID:26435876

  9. Medical personnel and patient dosimetry during coronary angiography and intervention

    NASA Astrophysics Data System (ADS)

    Efstathopoulos, Efstathios P.; Makrygiannis, Stamatis S.; Kottou, Sofia; Karvouni, Evangelia; Giazitzoglou, Eleftherios; Korovesis, Socrates; Tzanalaridou, Efthalia; Raptou, Panagiota D.; Katritsis, Demosthenes G.

    2003-09-01

    Percutaneous coronary interventions are associated with increased radiation exposure compared to most radiological examinations. This prospective study aimed at (1) measuring entrance doses for all in-room personnel, (2) performing an assessment of patient effective dose and intracoronary doses, (3) investigating the contribution of each projection to kerma-area product (KAP) and irradiation time, (4) comparing results with established DRL values in this clinical setting and (5) estimating the risk for fatal cancer to patients and operators. Measurements were performed during 40 consecutive procedures of coronary angiography (CA), half of which were followed by ad hoc coronary angioplasty (PTCA). KAP measurements were used for patients and thermoluminescent dosimetry for the in-room personnel. The mean KAP value per procedure for CA was 29 +/- 9 Gy cm2. Thirty four per cent of KAP was due to fluoroscopy, whereas the remainder (66%) was due to digital cine. Accordingly, the mean KAP value per PTCA procedure was 75 +/- 30 Gy cm2, and contribution of fluoroscopy is 57%. Effective dose per year was estimated to be 0.04-0.05 mSv y-1 for the primary operator, and 0.03-0.04 mSv y-1 for those assisting. Corresponding measurements for radiographer and nurse were below detectable level, implying minimal radiation hazards for them. Regarding radiation exposure, coronary intervention is considered a quite safe procedure for both patients and personnel in laboratories with modern equipment and experienced operators as long as standard safety precautions are considered. Exposure optimization though should be constantly sought through continuous review of procedures.

  10. Carotid artery stenting with double cerebral embolic protection in asymptomatic patients - a diffusion-weighted MRI controlled study.

    PubMed

    Vuruskan, Ertan; Saracoglu, Erhan; Ergun, Ugur; Poyraz, Fatih; Duzen, İrfan Veysel

    2017-01-01

    The aim of this study was to compare the simultaneous double-protection method (proximal balloon plus distal filter) with distal-filter protection or proximal-balloon protection alone in asymptomatic patients during carotid artery stenting. 119 consecutive patients were investigated for carotid artery stentings in the extracranial internal carotid artery with the use of distal filters (n = 41, 34.4 %), proximal balloon (MoMa) protection (n = 40, 33.6 %) or double protection (n = 38, 31.9 %). Magnetic resonance imaging (MRI) was performed on all patients before the procedure, and control diffusion-weighted MRI (DW-MRI) was obtained within 24-48 h after the procedure. Procedural data, complications, success rate, major adverse cardiovascular events, and MRI findings were collected. New cerebral high-intensity (HI) lesions were observed in 47 (39.4 %) patients. HI lesions were observed in 22 (53.6 %), 15 (37.5 %), and 10 (26.3 %) of the patients with distal filters, proximal protection, and double protection, respectively (p = 0.004). The average number of HI lesions on DW-MRI was 1.80 in the distal-filter group, 0.90 in the proximal-balloon group, and 0.55 in the double-protection group (p < 0.001). Procedure and fluoroscopy times were slightly longer in the double-protection group compared to the distal- or proximal-protection groups (p = 0.001). The double (proximal plus distal) cerebral embolic protection technique is safe and effective for minimizing the risk of cerebral embolization, even in patients with asymptomatic carotid artery stenosis, despite slightly longer procedure and fluoroscopy times.
.

  11. Face and construct validity of a computer-based virtual reality simulator for ERCP.

    PubMed

    Bittner, James G; Mellinger, John D; Imam, Toufic; Schade, Robert R; Macfadyen, Bruce V

    2010-02-01

    Currently, little evidence supports computer-based simulation for ERCP training. To determine face and construct validity of a computer-based simulator for ERCP and assess its perceived utility as a training tool. Novice and expert endoscopists completed 2 simulated ERCP cases by using the GI Mentor II. Virtual Education and Surgical Simulation Laboratory, Medical College of Georgia. Outcomes included times to complete the procedure, reach the papilla, and use fluoroscopy; attempts to cannulate the papilla, pancreatic duct, and common bile duct; and number of contrast injections and complications. Subjects assessed simulator graphics, procedural accuracy, difficulty, haptics, overall realism, and training potential. Only when performance data from cases A and B were combined did the GI Mentor II differentiate novices and experts based on times to complete the procedure, reach the papilla, and use fluoroscopy. Across skill levels, overall opinions were similar regarding graphics (moderately realistic), accuracy (similar to clinical ERCP), difficulty (similar to clinical ERCP), overall realism (moderately realistic), and haptics. Most participants (92%) claimed that the simulator has definite training potential or should be required for training. Small sample size, single institution. The GI Mentor II demonstrated construct validity for ERCP based on select metrics. Most subjects thought that the simulated graphics, procedural accuracy, and overall realism exhibit face validity. Subjects deemed it a useful training tool. Study repetition involving more participants and cases may help confirm results and establish the simulator's ability to differentiate skill levels based on ERCP-specific metrics.

  12. Radiation- and reference base-free navigation procedure for placement of instruments and implants: application to retrograde drilling of osteochondral lesions of the knee joint.

    PubMed

    Müller, Matthias; Gras, Florian; Marintschev, Ivan; Mückley, Thomas; Hofmann, Gunter O

    2009-01-01

    A novel, radiation- and reference base-free procedure for placement of navigated instruments and implants was developed and its practicability and precision in retrograde drillings evaluated in an experimental setting. Two different guidance techniques were used: One experimental group was operated on using the radiation- and reference base-free navigation technique (Fluoro Free), and the control group was operated on using standard fluoroscopy for guidance. For each group, 12 core decompressions were simulated by retrograde drillings in different artificial femurs following arthroscopic determination of the osteochondral lesions. The final guide-wire position was evaluated by postoperative CT analysis using vector calculation. High precision was achieved in both groups, but operating time was significantly reduced in the navigated group as compared to the control group. This was due to a 100% first-pass accuracy of drilling in the navigated group; in the control group a mean of 2.5 correction maneuvers per drilling were necessary. Additionally, the procedure was free of radiation in the navigated group, whereas 17.2 seconds of radiation exposure time were measured in the fluoroscopy-guided group. The developed Fluoro Free procedure is a promising and simplified approach to navigating different instruments as well as implants in relation to visually or tactilely placed pointers or objects without the need for radiation exposure or invasive fixation of a dynamic reference base in the bone.

  13. Percutaneous pedicle screw placement under single dimensional fluoroscopy with a designed pedicle finder-a technical note and case series.

    PubMed

    Tsuang, Fon-Yih; Chen, Chia-Hsien; Kuo, Yi-Jie; Tseng, Wei-Lung; Chen, Yuan-Shen; Lin, Chin-Jung; Liao, Chun-Jen; Lin, Feng-Huei; Chiang, Chang-Jung

    2017-09-01

    Minimally invasive spine surgery has become increasingly popular in clinical practice, and it offers patients the potential benefits of reduced blood loss, wound pain, and infection risk, and it also diminishes the loss of working time and length of hospital stay. However, surgeons require more intraoperative fluoroscopy and ionizing radiation exposure during minimally invasive spine surgery for localization, especially for guidance in instrumentation placement. In addition, computer navigation is not accessible in some facility-limited institutions. This study aimed to demonstrate a method for percutaneous screws placement using only the anterior-posterior (AP) trajectory of intraoperative fluoroscopy. A technical report (a retrospective and prospective case series) was carried out. Patients who received posterior fixation with percutaneous pedicle screws for thoracolumbar degenerative disease or trauma comprised the patient sample. We retrospectively reviewed the charts of consecutive 670 patients who received 4,072 pedicle screws between December 2010 and August 2015. Another case series study was conducted prospectively in three additional hospitals, and 88 consecutive patients with 413 pedicle screws were enrolled from February 2014 to July 2016. The fluoroscopy shot number and radiation dose were recorded. In the prospective study, 78 patients with 371 screws received computed tomography at 3 months postoperatively to evaluate the fusion condition and screw positions. In the retrospective series, the placement of a percutaneous screw required 5.1 shots (2-14, standard deviation [SD]=2.366) of AP fluoroscopy. One screw was revised because of a medialwall breach of the pedicle. In the prospective series, 5.8 shots (2-16, SD=2.669) were required forone percutaneous pedicle screw placement. There were two screws with a Grade 1 breach (8.6%), both at the lateral wall of the pedicle, out of 23 screws placed at the thoracic spine at T9-T12. Forthe lumbar and sacral areas, there were 15 Grade 1 breaches (4.3%), 1 Grade 2 breach (0.3%), and 1 Grade 3 breach (0.3%). No revision surgery was necessary. This method avoids lateral shots of fluoroscopy during screw placement and thus decreases the operation time and exposes surgeons to less radiation. At the same time, compared with the computer-navigated procedure, it is less facility-demanding, and provides satisfactory reliability and accuracy. Copyright © 2017 The Authors. Published by Elsevier Inc. All rights reserved.

  14. Accuracy of epicardial electroanatomic mapping and ablation of sustained ventricular tachycardia merged with heart CT scan in chronic Chagasic cardiomyopathy.

    PubMed

    Valdigem, Bruno Pereira; da Silva, Nilton José Carneiro; Dietrich, Cristiano Oliveira; Moreira, Dalmo; Sasdelli, Roberto; Pinto, Ibraim M; Cirenza, Claudio; de Paola, Angelo Amato Vincenzo

    2010-11-01

    As damage to coronary arteries is a potential complication of epicardial RF catheter ablation (EPRFCA), the procedure must be associated with coronary angiography. Chronic Chagasic cardiomiopathy (CCC) is a disease where epicardial VT are common. Eletroanatomic mapping merged with computed totmography (CT) scan data is a useful tool for mapping the endocardium, and its accuracy in guiding ablation on the epicardium was not adequately evaluated so far. Compare electronatomic map merged with Heart CT to fluoroscopy for epicardial ablation of CCC. Describe the distribution of the scars on CCC. We performed epicardial and endocardial mapping and ablation using CARTO XP V8 on eight patients and merged the map with coronary arteries CT scan using at least three landmarks. To compare the 3D image obtained with CARTO MERGE and the 2D fluoroscopic image obtained during the ablation procedure, we used computer graphic software (Inkscape™) in order to prove that the images were equivalent and to compare the distance between the catheter tip on fluoroscopy to catheter tip on 3D EA map. EPRFCA was successfully performed in all patients and they did not present recurrence for at least 3-month follow-up. The mean difference between the tip of the catheter on fluoroscopy and on the 3D model was 6.03 ± 2.09 mm. Scars were present in the epicardium and endocardium and most of patients presented with posterior wall scars and RV scar. The combination of electroanatomic map and CT coronary artery scan data is feasible and can be an important tool for EPRFCA in patients with CCC and VT.

  15. Reducing operator radiation exposure during cardiac resynchronization therapy.

    PubMed

    Brambilla, Marco; Occhetta, Eraldo; Ronconi, Martina; Plebani, Laura; Carriero, Alessandro; Marino, Paolo

    2010-12-01

    To quantify the reduction in equivalent dose at operator's hand that can be achieved by placement of a radiation-absorbing drape (RADPAD) during long-lasting cardiac resynchronization therapy (CRT) procedures. This is a prospective observational study that included 22 consecutive patients with drug-refractory heart failure who underwent implantation of a CRT device. The cases were randomly assigned to Group A (11 cases), performed without RADPAD, and to Group B (11 cases), performed using RADPAD. Dose equivalent at the examiner's hand was measured as H(p)(0.07) and as a time-adjusted H(p)(0.07) rate (mGy/min) with a direct reading dosimeter. The mean fluoroscopy time was 20.8 ± 7.7 min and the mean dose area product (DAP) was 118.6 ± 45.3 Gy cm(2). No significant differences were found between body mass index, fluoroscopy time, and DAP between patients examined with or without RADPAD. The correlation between the fluoroscopy time and the DAP was high (R(2) = 0.94, P < 0.001). Mean dose and dose rate measurement without the RADPAD at the finger and hand were H(p)(0.07) = 1.27 ± 0.47 mGy per procedure and H(p)(0.07) rate = 0.057 ± 0.011 mGy/min, respectively. The dosage was reduced with the RADPAD to H(p)(0.07) = 0.48 ± 0.20 (P < 0.05) and to H(p)(0.07) rate = 0.026 ± 0.008 (P < 0.001), respectively. A mean reduction of 54% in the equivalent dose rate to the operator's hand can be achieved with the use of RADPAD. The use of the RADPAD in CRT devices implantation will make unlikely the necessity of limiting the yearly number of implants for high volume operators.

  16. Ultrasound-guided versus fluoroscopy-guided sacroiliac joint intra-articular injections in the noninflammatory sacroiliac joint dysfunction: a prospective, randomized, single-blinded study.

    PubMed

    Jee, Haemi; Lee, Ji-Hae; Park, Ki Deok; Ahn, Jaeki; Park, Yongbum

    2014-02-01

    To compare the short-term effects and safety of ultrasound (US)-guided sacroiliac joint (SIJ) injections with fluoroscopy (FL)-guided SIJ injections in patients with noninflammatory SIJ dysfunction. Prospective, randomized controlled trial. University hospital. Patients (N=120) with noninflammatory sacroiliac arthritis were enrolled. All procedures were performed using an FL or US apparatus. Subjects were randomly assigned to either the FL or US group. Immediately after the SIJ injections, fluoroscopy was applied to verify the correct placement of the injected medication and intravascular injections. Treatment effects and functional improvement were compared at 2 and 12 weeks after the procedures. The verbal numeric pain scale and Oswestry Disability Index improved at 2 and 12 weeks after the injections without statistical significances between groups. Of 55 US-guided injections, 48 (87.3%) were successful and 7 (12.7%) were missed. The FL-guided SIJ approach exhibited a greater accuracy (98.2%) than the US-guided approach. Vascularization around the SIJ was seen in 34 of 55 patients. Among the 34 patients, 7 had vascularization inside the joint, 23 had vascularization around the joint, and 4 had vascularization both inside and around the joint. Three cases of intravascular injections occurred in the FL group. The US-guided approach may facilitate the identification and avoidance of the critical vessels around or within the SIJ. Function and pain relief significantly improved in both groups without significant differences between groups. The US-guided approach was shown to be as effective as the FL-guided approach in treatment effects. However, diagnostic application in the SIJ may be limited because of the significantly lower accuracy rate (87.3%). Copyright © 2014 American Congress of Rehabilitation Medicine. Published by Elsevier Inc. All rights reserved.

  17. CT fluoroscopy-guided core needle biopsy of anterior mediastinal masses.

    PubMed

    Iguchi, T; Hiraki, T; Matsui, Y; Fujiwara, H; Sakurai, J; Masaoka, Y; Uka, M; Tanaka, T; Gobara, H; Kanazawa, S

    2018-02-01

    To retrospectively evaluate the safety, diagnostic yield, and risk factors of diagnostic failure of computed tomography (CT) fluoroscopy-guided biopsies of anterior mediastinal masses. Biopsy procedures and results of anterior mediastinal masses in 71 patients (32 women/39 men; mean [±standard deviation] age, 53.8±20.0years; range, 14-88years) were analyzed. Final diagnoses were based on surgical outcomes, imaging findings, or clinical follow-up findings. The biopsy results were compared with the final diagnosis, and the biopsy procedures grouped by pathologic findings into diagnostic success and failure groups. Multiple putative risk factors for diagnostic failure were then assessed. Seventy-one biopsies (71 masses; mean size, 67.5±27.3mm; range 8.6-128.2mm) were analyzed. We identified 17 grade 1 and one grade 2 adverse events (25.4% overall) according to the National Cancer Institute Common Terminology Criteria for Adverse Events version 4.0. Sixty-nine biopsies (97.2%) provided samples fit for pathologic analysis. Diagnostic failure was found for eight (11.3%) masses; the 63 masses diagnosed successfully included thymic carcinoma (n=17), lung cancer (n=14), thymoma (n=12), malignant lymphoma (n=11), germ cell tumor (n=3), and others (n=6). Using a thinner needle (i.e., a 20-gauge needle) was the sole significant risk factor for diagnostic failure (P=0.039). CT fluoroscopy-guided biopsy of anterior mediastinal masses was safe and had a high diagnostic yield; however, using a thinner biopsy needle significantly increased the risk of a failed diagnosis. Copyright © 2017 Éditions françaises de radiologie. Published by Elsevier Masson SAS. All rights reserved.

  18. Application of newly developed Fluoro-QC software for image quality evaluation in cardiac X-ray systems.

    PubMed

    Oliveira, M; Lopez, G; Geambastiani, P; Ubeda, C

    2018-05-01

    A quality assurance (QA) program is a valuable tool for the continuous production of optimal quality images. The aim of this paper is to assess a newly developed automatic computer software for image quality (IR) evaluation in fluoroscopy X-ray systems. Test object images were acquired using one fluoroscopy system, Siemens Axiom Artis model (Siemens AG, Medical Solutions Erlangen, Germany). The software was developed as an ImageJ plugin. Two image quality parameters were assessed: high-contrast spatial resolution (HCSR) and signal-to-noise ratio (SNR). The time between manual and automatic image quality assessment procedures were compared. The paired t-test was used to assess the data. p Values of less than 0.05 were considered significant. The Fluoro-QC software generated faster IQ evaluation results (mean = 0.31 ± 0.08 min) than manual procedure (mean = 4.68 ± 0.09 min). The mean difference between techniques was 4.36 min. Discrepancies were identified in the region of interest (ROI) areas drawn manually with evidence of user dependence. The new software presented the results of two tests (HCSR = 3.06, SNR = 5.17) and also collected information from the DICOM header. Significant differences were not identified between manual and automatic measures of SNR (p value = 0.22) and HCRS (p value = 0.46). The Fluoro-QC software is a feasible, fast and free to use method for evaluating imaging quality parameters on fluoroscopy systems. Copyright © 2017 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.

  19. Multimode C-arm fluoroscopy, tomosynthesis, and cone-beam CT for image-guided interventions: from proof of principle to patient protocols

    NASA Astrophysics Data System (ADS)

    Siewerdsen, J. H.; Daly, M. J.; Bachar, G.; Moseley, D. J.; Bootsma, G.; Brock, K. K.; Ansell, S.; Wilson, G. A.; Chhabra, S.; Jaffray, D. A.; Irish, J. C.

    2007-03-01

    High-performance intraoperative imaging is essential to an ever-expanding scope of therapeutic procedures ranging from tumor surgery to interventional radiology. The need for precise visualization of bony and soft-tissue structures with minimal obstruction to the therapy setup presents challenges and opportunities in the development of novel imaging technologies specifically for image-guided procedures. Over the past ~5 years, a mobile C-arm has been modified in collaboration with Siemens Medical Solutions for 3D imaging. Based upon a Siemens PowerMobil, the device includes: a flat-panel detector (Varian PaxScan 4030CB); a motorized orbit; a system for geometric calibration; integration with real-time tracking and navigation (NDI Polaris); and a computer control system for multi-mode fluoroscopy, tomosynthesis, and cone-beam CT. Investigation of 3D imaging performance (noise-equivalent quanta), image quality (human observer studies), and image artifacts (scatter, truncation, and cone-beam artifacts) has driven the development of imaging techniques appropriate to a host of image-guided interventions. Multi-mode functionality presents a valuable spectrum of acquisition techniques: i.) fluoroscopy for real-time 2D guidance; ii.) limited-angle tomosynthesis for fast 3D imaging (e.g., ~10 sec acquisition of coronal slices containing the surgical target); and iii.) fully 3D cone-beam CT (e.g., ~30-60 sec acquisition providing bony and soft-tissue visualization across the field of view). Phantom and cadaver studies clearly indicate the potential for improved surgical performance - up to a factor of 2 increase in challenging surgical target excisions. The C-arm system is currently being deployed in patient protocols ranging from brachytherapy to chest, breast, spine, and head and neck surgery.

  20. Efficacy of a novel IGS system in atrial septal defect repair

    NASA Astrophysics Data System (ADS)

    Mefleh, Fuad N.; Baker, G. Hamilton; Kwartowitz, David M.

    2013-03-01

    Congenital heart disease occurs in 107.6 out of 10,000 live births, with Atrial Septal Defects (ASD) accounting for 10% of these conditions. Historically, ASDs were treated with open heart surgery using cardiopulmonary bypass, allowing a patch to be sewn over the defect. In 1976, King et al. demonstrated use of a transcatheter occlusion procedure, thus reducing the invasiveness of ASD repair. Localization during these catheter based procedures traditionally has relied on bi-plane fluoroscopy; more recently trans-esophageal echocardiography (TEE) and intra-cardiac echocardiography (ICE) have been used to navigate these procedures. Although there is a high success rate using the transcatheter occlusion procedure, fluoroscopy poses radiation dose risk to both patient and clinician. The impact of this dose to the patients is important as many of those undergoing this procedure are children, who have an increased risk associated with radiation exposure. Their longer life expectancy than adults provides a larger window of opportunity for expressing the damaging effects of ionizing radiation. In addition, epidemiologic studies of exposed populations have demonstrated that children are considerably more sensitive to the carcinogenic effects radiation. Image-guided surgery (IGS) uses pre-operative and intra-operative images to guide surgery or an interventional procedure. Central to every IGS system is a software application capable of processing and displaying patient images, registration between multiple coordinate systems, and interfacing with a tool tracking system. We have developed a novel image-guided surgery framework called Kit for Navigation by Image Focused Exploration (KNIFE). In this work we assess the efficacy of this image-guided navigation system for ASD repair using a series of mock clinical experiments designed to simulate ASD repair device deployment.

  1. A novel radiation protection drape reduces radiation exposure during fluoroscopy guided electrophysiology procedures.

    PubMed

    Germano, Joseph J; Day, Gina; Gregorious, David; Natarajan, Venkataraman; Cohen, Todd

    2005-09-01

    The purpose of this study was to evaluate a novel disposable lead-free radiation protection drape for decreasing radiation scatter during electrophysiology procedures. In recent years, there has been an exponential increase in the number of electrophysiology (EP) procedures exposing patients, operators and laboratory staff to higher radiation doses. The RADPAD was positioned slightly lateral to the incision site for pectoral device implants and superior to the femoral vein during electrophysiology studies. Each patient served as their own control and dosimetric measurements were obtained at the examiner's elbow and hand. Radiation badge readings for the operator were obtained three months prior to RADPAD use and three months after introduction. Radiation dosimetry was obtained in twenty patients: 7 electrophysiology studies, 6 pacemakers, 5 catheter ablations, and 2 implantable cardioverter-defibrillators. Eleven women and nine men with a mean age of 63 +/- 4 years had an average fluoroscopy time of 2.5 +/- 0.42 minutes per case. Mean dosimetric measurements at the hand were reduced from 141.38 +/- 24.67 to 48.63 +/- 9.02 milliroentgen (mR) per hour using the protective drape (63% reduction; p < 0.0001). Measurements at the elbow were reduced from 78.78 +/- 7.95 mR per hour to 34.50 +/- 4.18 mR per hour using the drape (55% reduction; p < 0.0001). Badge readings for three months prior to drape introduction averaged 2.45 mR per procedure versus 1.54 mR per procedure for 3 months post-initiation (37% reduction). The use of a novel radiation protection surgical drape can significantly reduce scatter radiation exposure to staff and operators during a variety of EP procedures.

  2. Deformable 3D-2D registration for CT and its application to low dose tomographic fluoroscopy

    NASA Astrophysics Data System (ADS)

    Flach, Barbara; Brehm, Marcus; Sawall, Stefan; Kachelrieß, Marc

    2014-12-01

    Many applications in medical imaging include image registration for matching of images from the same or different modalities. In the case of full data sampling, the respective reconstructed images are usually of such a good image quality that standard deformable volume-to-volume (3D-3D) registration approaches can be applied. But research in temporal-correlated image reconstruction and dose reductions increases the number of cases where rawdata are available from only few projection angles. Here, deteriorated image quality leads to non-acceptable deformable volume-to-volume registration results. Therefore a registration approach is required that is robust against a decreasing number of projections defining the target position. We propose a deformable volume-to-rawdata (3D-2D) registration method that aims at finding a displacement vector field maximizing the alignment of a CT volume and the acquired rawdata based on the sum of squared differences in rawdata domain. The registration is constrained by a regularization term in accordance with a fluid-based diffusion. Both cost function components, the rawdata fidelity and the regularization term, are optimized in an alternating manner. The matching criterion is optimized by a conjugate gradient descent for nonlinear functions, while the regularization is realized by convolution of the vector fields with Gaussian kernels. We validate the proposed method and compare it to the demons algorithm, a well-known 3D-3D registration method. The comparison is done for a range of 4-60 target projections using datasets from low dose tomographic fluoroscopy as an application example. The results show a high correlation to the ground truth target position without introducing artifacts even in the case of very few projections. In particular the matching in the rawdata domain is improved compared to the 3D-3D registration for the investigated range. The proposed volume-to-rawdata registration increases the robustness regarding sparse rawdata and provides more stable results than volume-to-volume approaches. By applying the proposed registration approach to low dose tomographic fluoroscopy it is possible to improve the temporal resolution and thus to increase the robustness of low dose tomographic fluoroscopy.

  3. Simulator-Based Angiography and Endovascular Neurosurgery Curriculum: A Longitudinal Evaluation of Performance Following Simulator-Based Angiography Training.

    PubMed

    Pannell, J Scott; Santiago-Dieppa, David R; Wali, Arvin R; Hirshman, Brian R; Steinberg, Jeffrey A; Cheung, Vincent J; Oveisi, David; Hallstrom, Jon; Khalessi, Alexander A

    2016-08-29

    This study establishes performance metrics for angiography and neuroendovascular surgery procedures based on longitudinal improvement in individual trainees with differing levels of training and experience. Over the course of 30 days, five trainees performed 10 diagnostic angiograms, coiled 10 carotid terminus aneurysms in the setting of subarachnoid hemorrhage, and performed 10 left middle cerebral artery embolectomies on a Simbionix Angio Mentor™ simulator. All procedures were nonconsecutive. Total procedure time, fluoroscopy time, contrast dose, heart rate, blood pressures, medications administered, packing densities, the number of coils used, and the number of stent-retriever passes were recorded. Image quality was rated, and the absolute value of technically unsafe events was recorded. The trainees' device selection, macrovascular access, microvascular access, clinical management, and the overall performance of the trainee was rated during each procedure based on a traditional Likert scale score of 1=fail, 2=poor, 3=satisfactory, 4=good, and 5=excellent. These ordinal values correspond with published assessment scales on surgical technique. After performing five diagnostic angiograms and five embolectomies, all participants demonstrated marked decreases in procedure time, fluoroscopy doses, contrast doses, and adverse technical events; marked improvements in image quality, device selection, access scores, and overall technical performance were additionally observed (p < 0.05). Similarly, trainees demonstrated marked improvement in technical performance and clinical management after five coiling procedures (p < 0.05). However, trainees with less prior experience deploying coils continued to experience intra-procedural ruptures up to the eighth embolization procedure; this observation likely corresponded with less tactile procedural experience to an exertion of greater force than appropriate for coil placement. Trainees across all levels of training and prior experience demonstrated a significant performance improvement after completion of our simulator curriculum consisting of five diagnostic angiograms, five embolectomy cases, and 10 aneurysm coil embolizations.

  4. Transthoracic Adrenal Biopsy Procedure Using Artificial Carbon Dioxide Pneumothorax as Outpatient Procedure

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Favelier, Sylvain; Guiu, Severine; Cherblanc, Violaine

    2013-08-01

    Many routes have been described for percutaneous adrenal gland biopsy. They require either a complex non-axial path or a long hydrodissection or even pass through an organ thereby increasing complications. We describe here an approach using an artificially-induced carbon dioxide (CO{sub 2}) pneumothorax, performed as an outpatient procedure in a 57-year-old woman. Under local anaesthesia, 200 ml of CO{sub 2} was injected in the pleural space through a Veress needle under computed tomography fluoroscopy, to clear the lung parenchyma from the biopsy route. Using this technique, transthoracic adrenal biopsy can be performed under simple local anaesthesia as an safely outpatientmore » procedure.« less

  5. Detection and correction of patient movement in prostate brachytherapy seed reconstruction

    NASA Astrophysics Data System (ADS)

    Lam, Steve T.; Cho, Paul S.; Marks, Robert J., II; Narayanan, Sreeram

    2005-05-01

    Intraoperative dosimetry of prostate brachytherapy can help optimize the dose distribution and potentially improve clinical outcome. Evaluation of dose distribution during the seed implant procedure requires the knowledge of 3D seed coordinates. Fluoroscopy-based seed localization is a viable option. From three x-ray projections obtained at different gantry angles, 3D seed positions can be determined. However, when local anaesthesia is used for prostate brachytherapy, the patient movement during fluoroscopy image capture becomes a practical problem. If uncorrected, the errors introduced by patient motion between image captures would cause seed mismatches. Subsequently, the seed reconstruction algorithm would either fail to reconstruct or yield erroneous results. We have developed an algorithm that permits detection and correction of patient movement that may occur between fluoroscopy image captures. The patient movement is decomposed into translational shifts along the tabletop and rotation about an axis perpendicular to the tabletop. The property of spatial invariance of the co-planar imaging geometry is used for lateral movement correction. Cranio-caudal movement is corrected by analysing the perspective invariance along the x-ray axis. Rotation is estimated by an iterative method. The method can detect and correct for the range of patient movement commonly seen in the clinical environment. The algorithm has been implemented for routine clinical use as the preprocessing step for seed reconstruction.

  6. Remote magnetic navigation for mapping and ablating right ventricular outflow tract tachycardia.

    PubMed

    Thornton, Andrew S; Jordaens, Luc J

    2006-06-01

    Navigation, mapping, and ablation in the right ventricular outflow tract (RVOT) can be difficult. Catheter navigation using external magnetic fields may allow more accurate mapping and ablation. The purpose of this study was to assess the feasibility of RVOT tachycardia ablation using remote magnetic navigation. Mapping and ablation were performed in eight patients with outflow tract ventricular arrhythmias. Tachycardia mapping was undertaken with a 64-polar basket catheter, followed by remote activation and pace-mapping using a magnetically enabled catheter. The area of interest was localized on the basket catheter in seven patients in whom an RVOT arrhythmia was identified. Remote navigation of the magnetic catheter to this area was followed by pace-mapping. Ablation was performed at the site of perfect pace-mapping, with earliest activation if possible. Acute success was achieved in all patients (median four applications). Median procedural time was 144 minutes, with 13.4 minutes of patient fluoroscopy time and 3.8 minutes of physician fluoroscopy time. No complications occurred. One recurrence occurred during follow-up (mean 366 days). RVOT tachycardias can be mapped and ablated using remote magnetic navigation, initially guided by a basket catheter. Precise activation and pace-mapping are possible. Remote magnetic navigation permitted low fluoroscopy exposure for the physician. Long-term results are promising.

  7. Fluoroscopically guided transforaminal epidural steroid injections at a quaternary-care teaching institution: effect of trainee involvement and patient body mass index on fluoroscopy time and patient dose.

    PubMed

    Tiegs-Heiden, C A; Murthy, N S; Geske, J R; Diehn, F E; Schueler, B A; Wald, J T; Kaufmann, T J; Lehman, V T; Carr, C M; Amrami, K K; Morris, J M; Thielen, K R; Maus, T P

    2016-01-01

    To investigate whether there are differences in fluoroscopy time and patient dose for fluoroscopically guided lumbar transforaminal epidural steroid injections (TFESIs) performed by staff radiologists versus with trainees and to evaluate the effect of patient body mass index (BMI) on fluoroscopy time and patient dose, including their interactions with other variables. Single-level lumbar TFESIs (n=1844) between 1 January 2011 and 31 December 2013 were reviewed. Fluoroscopy time, reference point air kerma (Ka,r), and kerma area product (KAP) were recorded. BMI and trainee involvement were examined as predictors of fluoroscopy time, Ka,r, and KAP in models adjusted for age and gender in multivariable linear models. Stratified models of BMI groups by trainee presence were performed. Increased age was the only significant predictor of increased fluoroscopy time (p<0.0001). Ka,r and KAP were significantly higher in patients with a higher BMI (p<0.0001 and p=0.0009). When stratified by BMI, longer fluoroscopy time predicted increased Ka,r and KAP in all groups (p<0.0001). Trainee involvement was not a statistically significant predictor of fluoroscopy time or Ka,r in any BMI category. KAP was lower with trainees in the overweight group (p=0.0009) and higher in male patients for all BMI categories (p<0.02). Trainee involvement did not result in increased fluoroscopy time or patient dose. BMI did not affect fluoroscopy time; however, overweight and obese patients received significantly higher Ka,r and KAP. Male patients received a higher KAP in all BMI categories. Limiting fluoroscopy time and good collimation practices should be reinforced in these patients. Copyright © 2015 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.

  8. The modern treatment of oesophageal strictures using the Eder-Puestow dilators.

    PubMed

    Borgeskov, S; Struve-Christensen, E

    1978-01-01

    The greatest safety in bouginage of narrow and twisted oesophageal strictures is obtained by employing a flexible oesophagoscope via which a guiding probe is introduced by Puestow's method. The actual bougie may then be introduced over this guide without risk of perforating the oesophagus. When performed under TV-fluoroscopy, this procedure gains added safety. 108 dilations were performed without complications.

  9. The Usefulness of Virtual Fluoroscopic Preprocedural Planning During Percutaneous Transhepatic Biliary Drainage

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kinoshita, Mitsuhiro, E-mail: kinoshita.3216@tokushima-u.ac.jp; Shirono, Ryozo; Takechi, Katsuya

    Purpose To retrospectively evaluate the usefulness of virtual fluoroscopic preprocedural planning (VFPP) in the percutaneous transhepatic biliary drainage (PTBD) procedure.Materials and MethodsTwenty-two patients who were treated by PTBD were included in this study. Twelve patients were treated using PTBD intraoperative referencing coronal computed tomography (CT) images (i.e., coronal CT group), and ten patients were treated using PTBD intraoperative referencing VFPP images (i.e., VFPP group). To analyze the effect of the intraoperative referencing VFPP image, the VFPP group was retrospectively compared with the coronal CT group.ResultsThe characteristics of both patient groups were not statistically significantly different. There were no significant differencesmore » in the targeted bile duct, diameter and depth of the target bile, breath-holding ability, number of targeted bile duct puncture attempts, change in the targeted bile duct, and exchange of the drainage catheter. However, the X-ray fluoroscopy time and the procedure time were significantly shorter in the VFPP group than in the coronal CT group (196 vs. 334 s, P < 0.05; and 16.0 vs. 27.2 min, P < 0.05).ConclusionIntraoperative referencing using the VFPP imaging in PTBD intuitively can be a useful tool for better localization of the guidewire in the bile duct and thereby shorten the X-ray fluoroscopy time and procedure time while minimizing radiation exposure and complications.« less

  10. Transesophageal echocardiography and intracardiac echocardiography differently predict potential technical challenges or failures of interatrial shunts catheter-based closure.

    PubMed

    Rigatelli, Gianluca; Rigateli, Gianluca; Cardaioli, Paolo; Braggion, Gabriele; Aggio, Silvio; Giordan, Massimo; Magro, Beatrice; Nascimben, Alberto; Favaro, Alberto; Roncon, Loris; Rincon, Loris

    2007-02-01

    We sought to prospectively assess the role of transesophageal (TEE) and intracardiac echocardiography (ICE) in detecting potential technical difficulties or failures in patients submitted to interatrial shunts percutaneous closure. We prospectively enrolled 46 consecutive patients (mean age 35+/-28, 8 years, 30 female) referred to our center for catheter-based closure of interatrial shunts. All patients were screened with TEE before the intervention. Patients who met the inclusion criteria underwent ICE study before the closure attempt (40 patients). TEE detected potential technical difficulties in 22.5% (9/40) patients, whereas ICE detected technical difficulties in 32.5% (13/40 patients). In patients with positive TEE/ICE the procedural success (92.4% versus 100% and, P = ns) and follow-up failure rate (7.7% versus 0%, P = ns) were similar to patients with negative TEE/ICE, whereas the fluoroscopy time (7 +/- 1.2 versus 5 +/- 0.7 minutes, P < 0.03), the procedural time (41 +/- 4.1 versus 30 +/- 8.2 minutes, P +/- 0.03), and technical difficulties rate (23.1% versus 0%, P = 0.013) were higher. Differences between ICE and TEE in the evaluation of rims, measurement of ASD or fossa ovalis, and detection of venous valve and embryonic septal membrane remnants impacted on technical challenges and on procedural and fluoroscopy times but did not influence the success rate and follow-up failure rate.

  11. Decreased Radiation Exposure and Increased Efficacy in Extracorporeal Lithotripsy Using a New Ultrasound Stone Locking System.

    PubMed

    Abid, Nadia; Ravier, Emmanuel; Promeyrat, Xavier; Codas, Ricardo; Fehri, Hakim Fassi; Crouzet, Sebastien; Martin, Xavier

    2015-11-01

    To compare fluoroscopy duration, radiation dose, and efficacy of two ultrasound stone localization systems during extracorporeal shockwave lithotripsy (SWL) treatment. Monocentric prospective data were obtained from patients consecutively treated for renal stones using the Sonolith(®) i-sys (EDAP TMS) lithotripter, with fluoroscopy combined with ultrasound localization using an "outline" Automatic Ultrasound Positioning Support (AUPS) (group A), or the "free-line" Visio-Track (VT) (EDAP-TMS) hand-held three-dimensional ultrasound stone locking system (group B). Efficacy rate was defined as the within-groups proportion stone free or with partial stone fragmentation not needing additional procedures. Statistical analysis used Pearson chi-square tests for categoric variables, nonparametric Mann-Whitney tests for continuous variables, and linear regression for operator learning curve with VT. Continuous variables were reported as median (range) values. Patients in group A (n=73) and group B (n=81) were comparable in baseline characteristics (age, kidney stone size, others) and in SWL application (duration, number of shocks, energy [Joules]). During SWL, the median (range) duration (seconds) of radiation exposure was 159.5 (0-690) in group A and 3.5 (0-478) in group B (P<0.001) and irradiation dose (mGy.cm(2)), 10598 (0-54843) in group A and 163 (0-13926) in group B (P<0.001). Fluoroscopy time significantly decreased with operator experience using VT. The efficacy rate was 54.5% in group A and 79.5% in group B (P=0.001). VT significantly reduced fluoroscopy use during SWL and the duration and dose of patient exposure to ionizing radiation. Stone treatment efficacy was significantly greater with VT mainly because of a better real-time monitoring of the stone.

  12. Use of fluoroscopy-guided wire manipulation and/or laparoscopic surgery in the repair of malfunctioning peritoneal dialysis catheters.

    PubMed

    Kim, Hee Jin; Lee, Tae Won; Ihm, Chun Gyoo; Kim, Myung Jae

    2002-01-01

    Peritoneal catheter is the lifeline for the continuous ambulatory peritoneal dialysis (CAPD) patients. Over the years, obstruction or displacement of the CAPD catheter has been one of the common complications of CAPD. Fluoroscopy-guided wire manipulation or laparoscopic surgery has been developed to manage outflow obstruction. We analyzed the catheter outcome of fluoroscopy-guided wire manipulation or laparoscopic surgery to determine the ultimate benefit of these procedures. From June 1996 to August 2000, catheter complications were manipulated in 24 patients. Eleven (46%) of these patients were initially managed by guide wire under fluoroscopic control. The remaining 13 (54%) patients were manipulated by laparoscopic surgery. A successful outcome was defined as maintained normal peritoneal catheter function at 30 days after the manipulations. Among the catheters manipulated, 18 (75%) were inserted by nephrologist and 6 (25%) by surgeons at the initiation of CAPD. Tenckhoff double-cuff peritoneal catheters were inserted to all patients. The time elapsed between catheter insertion and manipulation varied from 1 to 60 days with a mean of 11 days. The primary causes of catheter malfunction were kinking in 1 case, omental wrapping with adhesions in 9 cases, and catheter displacements in the remaining 14 cases. Thirty-day catheter function was achieved in 50% (12/24) of initial catheter manipulations, with guide wire under fluoroscopic control (46%, 5/11) and laparoscopic surgery (54%, 7/13). Overall success rate of repeated manipulation was 71% (17 of 24). The successful outcome in repairing of the malfunctioning CAPD catheters could be increased by the combination of fluoroscopy-guided wire manipulation and laparoscopic surgery. Copyright 2002 S. Karger AG, Basel

  13. Sports-Related Groin Pain Secondary to Symphysis Pubis Disorders: Correlation Between MRI Findings and Outcome After Fluoroscopy-Guided Injection of Steroid and Local Anesthetic.

    PubMed

    Byrne, Caoimhe A; Bowden, Dermot J; Alkhayat, Abdullah; Kavanagh, Eoin C; Eustace, Stephen J

    2017-08-01

    The objective of our study was to correlate patterns of injury on preprocedural MRI with outcome after targeted fluoroscopy-guided steroid and local anesthetic injection of the symphysis pubis and its muscular attachments in a group of athletes with chronic groin pain. Forty-five patients with chronic sports-related groin pain underwent MRI of the pelvis and a targeted fluoroscopy-guided symphyseal corticosteroid and local anesthetic injection. Preprocedural MRI was reviewed. The presence or absence of a "superior cleft" sign (i.e., rectus abdominis-adductor longus attachment microtearing), "secondary cleft" sign (i.e., short adductor [gracilis, adductor brevis, and pectineus muscles] attachment microtearing), osteitis pubis, and extrasymphyseal pelvic abnormalities was recorded. Patients were followed up a mean time of 23 months after the procedure, and outcome was recorded. Correlation was made between preprocedural MRI findings and outcome. Forty-two percent of the patients had an isolated superior cleft sign, 7% had an isolated secondary cleft sign, and 11% had isolated osteitis pubis. Thirty-one percent of patients had a more complex injury, and 9% had a normal symphysis pubis. Overall, 89% of the patients experienced an improvement in symptoms. The response was sustained after a minimum of 6 months in 58% of the patients. The presence of the superior cleft sign was more frequently associated with a complete recovery. Fluoroscopy-guided corticosteroid symphyseal injection is a safe and effective treatment of sports-related groin pain. It is more frequently associated with a complete recovery in patients who display an isolated superior cleft sign on MRI. MRI not only is useful in characterizing groin injuries but also may be helpful in predicting response to therapeutic injection.

  14. Fluoroscopy of spontaneous breathing is more sensitive than phrenic nerve stimulation for detection of right phrenic nerve injury during cryoballoon ablation of atrial fibrillation.

    PubMed

    Linhart, Markus; Nielson, Annika; Andrié, René P; Mittmann-Braun, Erica L; Stöckigt, Florian; Kreuz, Jens; Nickenig, Georg; Schrickel, Jan W; Lickfett, Lars M

    2014-08-01

    Right phrenic nerve palsy (PNP) is a typical complication of cryoballoon ablation of the right-sided pulmonary veins (PVs). Phrenic nerve function can be monitored by palpating the abdomen during phrenic nerve pacing from the superior vena cava (SVC pacing) or by fluoroscopy of spontaneous breathing. We sought to compare the sensitivity of these 2 techniques during cryoballoon ablation for detection of PNP. A total of 133 patients undergoing cryoballoon ablation were monitored with both SVC pacing and fluoroscopy of spontaneous breathing during ablation of the right superior PV. PNP occurred in 27/133 patients (20.0%). Most patients (89%) had spontaneous recovery of phrenic nerve function at the end of the procedure or on the following day. Three patients were discharged with persistent PNP. All PNP were detected first by fluoroscopic observation of diaphragm movement during spontaneous breathing, while diaphragm could still be stimulated by SVC pacing. In patients with no recovery until discharge, PNP occurred at a significantly earlier time (86 ± 34 seconds vs. 296 ± 159 seconds, P < 0.001). No recovery occurred in 2/4 patients who were ablated with a 23 mm cryoballoon as opposed to 1/23 patients with a 28 mm cryoballoon (P = 0.049). Fluoroscopic assessment of diaphragm movement during spontaneous breathing is more sensitive for detection PNP as compared to SVC pacing. PNP as assessed by fluoroscopy is frequent (20.0%) and carries a high rate of recovery (89%) until discharge. Early onset of PNP and use of 23 mm cryoballoon are associated with PNP persisting beyond hospital discharge. © 2014 Wiley Periodicals, Inc.

  15. Outcomes of repeat catheter ablation using magnetic navigation or conventional ablation.

    PubMed

    Akca, Ferdi; Theuns, Dominic A M J; Abkenari, Lara Dabiri; de Groot, Natasja M S; Jordaens, Luc; Szili-Torok, Tamas

    2013-10-01

    After initial catheter ablation, repeat procedures could be necessary. This study evaluates the efficacy of the magnetic navigation system (MNS) in repeat catheter ablation as compared with manual conventional techniques (MANs). The results of 163 repeat ablation procedures were analysed. Ablations were performed either using MNS (n = 84) or conventional manual ablation (n = 79). Procedures were divided into four groups based on the technique used during the initial and repeat ablation procedure: MAN-MAN (n = 66), MAN-MNS (n = 31), MNS-MNS (n = 53), and MNS-MAN (n = 13). Three subgroups were analysed: supraventricular tachycardias (SVTs, n = 68), atrial fibrillation (AF, n = 67), and ventricular tachycardias (VT, n = 28). Recurrences were assessed during 19 ± 11 months follow-up. Overall, repeat procedures using MNS were successful in 89.0% as compared with 96.2% in the MAN group (P = ns). The overall recurrence rate was significantly lower using MNS (25.0 vs. 41.4%, P = 0.045). Acute success and recurrence rates for the MAN-MAN, MAN-MNS, MNS-MNS, and MNS-MAN groups were comparable. For the SVT subgroup a higher acute success rate was achieved using MAN (87.9 vs. 100.0%, P = 0.049). The use of MNS for SVT is associated with longer procedure times (205 ± 82 vs. 172 ± 69 min, P = 0.040). For AF procedure and fluoroscopy times were longer (257 ± 72 vs. 185 ± 64, P = 0.001; 59.5 ± 19.3 vs. 41.1 ± 18.3 min, P < 0.001). Less fluoroscopy was used for MNS-guided VT procedures (22.8 ± 14.7 vs. 41.2 ± 10.9, P = 0.011). Our data suggest that overall MNS is comparable with MAN in acute success after repeat catheter ablation. However, MNS is related to fewer recurrences as compared with MAN.

  16. Demonstration of the Safety and Feasibility of Robotically Assisted Percutaneous Coronary Intervention in Complex Coronary Lesions: Results of the CORA-PCI Study (Complex Robotically Assisted Percutaneous Coronary Intervention).

    PubMed

    Mahmud, Ehtisham; Naghi, Jesse; Ang, Lawrence; Harrison, Jonathan; Behnamfar, Omid; Pourdjabbar, Ali; Reeves, Ryan; Patel, Mitul

    2017-07-10

    The aims of this study were to evaluate the feasibility and technical success of robotically assisted percutaneous coronary intervention (R-PCI) for the treatment of coronary artery disease (CAD) in clinical practice, especially in complex lesions, and to determine the safety and clinical success of R-PCI compared with manual percutaneous coronary intervention (M-PCI). R-PCI is safe and feasible for simple coronary lesions. The utility of R-PCI for complex coronary lesions is unknown. All consecutive PCI procedures performed robotically (study group) or manually (control group) over 18 months were included. R-PCI technical success, defined as the completion of the procedure robotically or with partial manual assistance and without a major adverse cardiovascular event, was determined. Procedures ineligible for R-PCI (i.e., atherectomy, planned 2-stent strategy for bifurcation lesion, chronic total occlusion requiring hybrid approach) were excluded for analysis from the M-PCI group. Clinical success, defined as completion of the PCI procedure without a major adverse cardiovascular event, procedure time, stent use, and fluoroscopy time were compared between groups. A total of 315 patients (mean age 67.7 ± 11.8 years; 78% men) underwent 334 PCI procedures (108 R-PCIs, 157 lesions, 78.3% type B2/C; 226 M-PCIs, 336 lesions, 68.8% type B2/C). Technical success with R-PCI was 91.7% (rate of manual assistance 11.1%, rate of manual conversion 7.4%, rate of major adverse cardiovascular events 0.93%). Clinical success (99.1% with R-PCI vs. 99.1% with M-PCI; p = 1.00), stent use (stents per procedure 1.59 ± 0.79 with R-PCI vs. 1.54 ± 0.75 with M-PCI; p = 0.73), and fluoroscopy time (18.2 ± 10.4 min with R-PCI vs. 19.2 ± 11.4 min with M-PCI; p = 0.39) were similar between the groups, although procedure time was longer in the R-PCI group (44:30 ± 26:04 min:s vs. 36:34 ± 23:03 min:s; p = 0.002). Propensity-matched analysis confirmed that procedure time was longer in the robotic group (42:59 ± 26:14 min:s with R-PCI vs. 34:01 ± 17:14 min:s with M-PCI; p = 0.007), although clinical success remained similar (98.8% with R-PCI vs. 100% with M-PCI; p = 1.00). This study demonstrates the feasibility, safety, and high technical success of R-PCI for the treatment of complex coronary disease. Furthermore, comparable clinical outcomes, without an adverse effect on stent use or fluoroscopy time, were observed with R-PCI and M-PCI. Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

  17. Patient radiation doses in interventional cardiology in the U.S.: Advisory data sets and possible initial values for U.S. reference levels

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Miller, Donald L.; Hilohi, C. Michael; Spelic, David C.

    2012-10-15

    Purpose: To determine patient radiation doses from interventional cardiology procedures in the U.S and to suggest possible initial values for U.S. benchmarks for patient radiation dose from selected interventional cardiology procedures [fluoroscopically guided diagnostic cardiac catheterization and percutaneous coronary intervention (PCI)]. Methods: Patient radiation dose metrics were derived from analysis of data from the 2008 to 2009 Nationwide Evaluation of X-ray Trends (NEXT) survey of cardiac catheterization. This analysis used deidentified data and did not require review by an IRB. Data from 171 facilities in 30 states were analyzed. The distributions (percentiles) of radiation dose metrics were determined for diagnosticmore » cardiac catheterizations, PCI, and combined diagnostic and PCI procedures. Confidence intervals for these dose distributions were determined using bootstrap resampling. Results: Percentile distributions (advisory data sets) and possible preliminary U.S. reference levels (based on the 75th percentile of the dose distributions) are provided for cumulative air kerma at the reference point (K{sub a,r}), cumulative air kerma-area product (P{sub KA}), fluoroscopy time, and number of cine runs. Dose distributions are sufficiently detailed to permit dose audits as described in National Council on Radiation Protection and Measurements Report No. 168. Fluoroscopy times are consistent with those observed in European studies, but P{sub KA} is higher in the U.S. Conclusions: Sufficient data exist to suggest possible initial benchmarks for patient radiation dose for certain interventional cardiology procedures in the U.S. Our data suggest that patient radiation dose in these procedures is not optimized in U.S. practice.« less

  18. Ultrasound and Fluoroscopy-Guided Placement of Central Venous Ports via Internal Jugular Vein: Retrospective Analysis of 1254 Port Implantations at a Single Center

    PubMed Central

    Ahn, Se Jin; Chung, Jin Wook; An, Sang Bu; Yin, Yong Hu; Jae, Hwan Jun; Park, Jae Hyung

    2012-01-01

    Objective To assess the technical success and complication rates of the radiologic placement of central venous ports via the internal jugular vein. Materials and Methods We retrospectively reviewed 1254 central venous ports implanted at our institution between August 2002 and October 2009. All procedures were guided by using ultrasound and fluoroscopy. Catheter maintenance days, technical success rates, peri-procedural, as well as early and late complication rates were evaluated based on the interventional radiologic reports and patient medical records. Results A total of 433386 catheter maintenance days (mean, 350 days; range 0-1165 days) were recorded. The technical success rate was 99.9% and a total of 61 complications occurred (5%), resulting in a post-procedural complication rate of 0.129 of 1000 catheter days. Among them, peri-procedural complications within 24 hours occurred in five patients (0.4%). There were 56 post-procedural complications including 24 (1.9%, 0.055 of 1000 catheter days) early and 32 (2.6%, 0.074 of 1000 catheter days) late complications including, infection (0.6%, 0.018 of 10000 catheter days), thrombotic malfunction (1.4%, 0.040 of 1000 catheter days), nonthrombotic malfunction (0.9%, 0.025 of 1000 catheter days), venous thrombosis (0.5%, 0.014 of 1000 catheter days), as well as wound problems (1.1%, 0.032 of 1000 catheter days). Thirty six CVPs (3%) were removed due to complications. Bloodstream infections and venous thrombosis were the two main adverse events prolonging hospitalization (mean 13 days and 5 days, respectively). Conclusion Radiologic placement of a central venous port via the internal jugular vein is safe and efficient as evidenced by its high technical success rate and a very low complication rate. PMID:22563269

  19. Time-Course Reduction in Patient Exposure to Radiation From Coronary Interventional Procedures: The Greater Paris Area Percutaneous Coronary Intervention Registry.

    PubMed

    Georges, Jean-Louis; Karam, Nicole; Tafflet, Muriel; Livarek, Bernard; Bataille, Sophie; Loyeau, Aurélie; Mapouata, Mireille; Benamer, Hakim; Caussin, Christophe; Garot, Philippe; Varenne, Olivier; Barbou, Franck; Teiger, Emmanuel; Funck, François; Karrillon, Gaëtan; Lambert, Yves; Spaulding, Christian; Jouven, Xavier

    2017-08-01

    The frequency of complex percutaneous coronary interventions (PCIs) has increased in the last few years, with a growing concern on the radiation dose received by the patients. Multicenter data from large unselected populations on patients' radiation doses during coronary angiography (CA) and PCI and temporal trends are lacking. This study sought to evaluate the temporal trends in patients' exposure to radiation from CA and PCI. Data were taken from the CARDIO-ARSIF registry that prospectively collects data on all CAs and PCIs performed in the 36 catheterization laboratories in the Greater Paris Area, the most populated regions in France with about 12 million inhabitants. Kerma area product and Fluoroscopy time from 152 684 consecutive CAs and 103 177 PCIs performed between 2009 and 2013 were analyzed. A continuous trend for a decrease in median [interquartile range] Kerma area product was observed, from 33 [19-55] Gy cm 2 in 2009 to 27 [16-44] Gy cm 2 in 2013 for CA ( P <0.0001), and from 73 [41-125] to 55 [31-91] Gy cm 2 for PCI ( P <0.0001). Time-course differences in Kerma area product remained highly significant after adjustment on Fluoroscopy time, PCI procedure complexity, change of x-ray equipment, and other patient- and procedure-related covariates. In a large patient population, a steady temporal decrease in patient radiation exposure during CA and PCI was noted between 2009 and 2013. Kerma area product reduction was consistent in all types of procedure and was independent of patient-related factors and PCI procedure complexity. © 2017 American Heart Association, Inc.

  20. Intra-operative localisation of thoracic spine level: a simple "'K'-wire in pedicle" technique.

    PubMed

    Thambiraj, Sathya; Quraishi, Nasir A

    2012-05-01

    To describe a simple and reliable method of intra-operative localisation of thoracic spine in a single surgical setting. Intra-operative localisation of thoracic spine levels can be difficult due to anatomical constraints, such as scapular shadow, patient's size and poor bone quality. This is particularly true in cases of thoracic discectomies in which the vertebral bodies appear normal. There are several methods described in recent literature to address this. Many of them require a separate procedure which was performed often the previous day. We report a technique which addresses the issue of localising thoracic level intra-operatively. After induction of general anaesthesia, the patient was placed prone and the pedicle of interest was identified using fluoroscopy. A K-wire was then inserted percutaneously into this pedicle under image guidance [confirmed in the antero-posterior (AP) and lateral views]. The wire was then cut close to the skin after bending it. The patient was now positioned laterally and the intended procedure performed through an anterior trans-thoracic approach. The 'K' wire was removed at the end of the procedure. We routinely used this technique in all our thoracic discectomies (four cases in 2 years). There were no intra-operative complications. This method is simple, avoids the patient undergoing two procedures and requires no more ability than placing an implant in the pedicle under fluoroscopy. Placing the 'K' wire into a fixed point like the pedicle facilitates rapid intra-operative viewing of the level of interest and is removed easily at the conclusion of surgery.

  1. "More may mean less... " the role for simulation-based medical education in the cardiac catheterization laboratory.

    PubMed

    Westerdahl, Daniel E; Henry, Timothy D

    2016-02-15

    Implementation of simulation-based medical education (SBME) can improve cardiovascular fellows' angiography skills and knowledge SBME focused on performing coronary angiography shortened procedure times and decreased the use of cine-fluoroscopy The ACGME mandate and SCAI's Simulation Committee recommendations suggest SBME will play an expanding and integral role in the field of cardiovascular medicine. © 2016 Wiley Periodicals, Inc.

  2. Remote magnetic with open-irrigated catheter vs. manual navigation for ablation of atrial fibrillation: a systematic review and meta-analysis.

    PubMed

    Proietti, Riccardo; Pecoraro, Valentina; Di Biase, Luigi; Natale, Andrea; Santangeli, Pasquale; Viecca, Maurizio; Sagone, Antonio; Galli, Alessio; Moja, Lorenzo; Tagliabue, Ludovica

    2013-09-01

    The aim of this study was to determine the efficacy and safety of remote magnetic navigation (RMN) with open-irrigated catheter vs. manual catheter navigation (MCN) in performing atrial fibrillation (AF) ablation. We searched in PubMed (1948-2013) and EMBASE (1974-2013) studies comparing RMN with MCN. Outcomes considered were AF recurrence (primary outcome), pulmonary vein isolation (PVI), procedural complications, and data on procedure's performance. Odds ratios (OR) and mean difference (MD) were extracted and pooled using a random-effect model. Confidence in the estimates of the obtained effects (quality of evidence) was assessed using the Grading of Recommendations Assessment, Development and Evaluation approach. We identified seven controlled trials, six non-randomized and one randomized, including a total of 941 patients. Studies were at high risk of bias. No difference was observed between RMN and MCN on AF recurrence [OR 1.18, 95% confidence interval (CI) 0.85 to 1.65, P = 0.32] or PVI (OR 0.41, 95% CI 0.11-1.47, P = 0.17). Remote magnetic navigation was associated with less peri-procedural complications (Peto OR 0.41, 95% CI 0.19-0.88, P = 0.02). Mean fluoroscopy time was reduced in RMN group (-22.22 min; 95% CI -42.48 to -1.96, P = 0.03), although the overall duration of the procedure was longer (60.91 min; 95% CI 31.17 to 90.65, P < 0.0001). In conclusion, RMN is not superior to MCN in achieving freedom from recurrent AF at mid-term follow-up or PVI. The procedure implies less peri-procedural complications, requires a shorter fluoroscopy time but a longer total procedural time. For the low quality of the available evidence, a proper designed randomized controlled trial could turn the direction and the effect of the dimensions explored.

  3. Efficacy and complications in the use of self-expanding colonic stents: an analysis of 15 years' experience.

    PubMed

    Samper Wamba, J D; Fernández Martínez, A; González Pastrana, L; López González, L; Balboa Arregui, Ó

    2015-01-01

    To analyze the efficacy and safety of the procedure for placing self-expanding stents in the colon. To evaluate the factors associated with complications. To analyze the dose of radiation delivered in the procedure. This was a retrospective descriptive study of 478 procedures done at a single center to place self-expanding metallic stents in the colon. A total of 423 nitinol stents and 79 stainless steel stents were placed. We included all colonic obstructions, of which 446 had malignant causes and 8 had benign causes. We excluded patients with intestinal perforation, severe colonic bleeding, short life expectancy, or lesions located less than 5 cm from the anus. We collected the dosimetric data and analyzed the technical success, clinical success, and complications during follow-up. The procedure was a technical success in 92.26% of cases (n=441) and a clinical success in 78.45% (n=375); complications occurred during follow-up in 18.5% of cases. Complications occurred more frequently with the stainless steel stents than with the nitinol stents (OR: 3.2; 95% CI: 1.8-5.7). The mean value of the dose area product was 35 Gy*cm(2). When instead of being done by the interventional radiologist working together with an endoscopist the procedure was done exclusively by the interventional radiologist, the time under fluoroscopy (p=0.001), dose area product (p=0.029), and kinetic energy released per unit mass (p=0.001) were greater. The procedure for placing self-expanding colonic stents is efficacious and safe with an acceptable rate of complications. The doses of radiation delivered were low, and the radiation doses and time under fluoroscopy were lower when the procedure was done together with an endoscopist. Copyright © 2014 SERAM. Published by Elsevier España, S.L.U. All rights reserved.

  4. A systematic review of the uses of fluoroscopy in dentistry.

    PubMed

    Uzbelger Feldman, Daniel; Yang, Jie; Susin, Cristiano

    2010-01-01

    To determine the quality of the evidence for the uses of fluoroscopy in dentistry. A systematic review using Ovid and MEDLINE was conducted to identify papers showing the uses of fluoroscopy in dentistry published between 1953 and September 2009. Human, animal and phantom/skull/mannequin studies on fluoroscopy with regard to its diagnostic value, research performance, and clinical and safety applications in dentistry were included in this analysis. Studies that were not in English, as well as those that employed fluoroscopy in dentistry without the use of image intensification, were excluded. Articles were evaluated, classified and graded by levels of evidence. Fifty-five out of 139 papers fulfilled the inclusion criteria. Amongst them, 19 were related to diagnosis, 15 to research, 12 to clinical and nine to safety applications. Fluoroscopy has contributed to nine different areas of dentistry. Also, it was used on 895 dental patients, 37 animals and 17 phantoms/skulls/mannequins. Two randomised controlled trials, two cohort studies, two case controls, 48 case reports and one expert opinion were found. Fluoroscopy with image intensification has been a useful, but not consistently used tool in dentistry for over 50 years. Several lines of evidence have shown fluoroscopy's diagnostic potential, research use, and clinical and safety applications in dentistry.

  5. Orthopedic surgeons' knowledge regarding risk of radiation exposition: a survey analysis.

    PubMed

    Tunçer, Nejat; Kuyucu, Ersin; Sayar, Şafak; Polat, Gökhan; Erdil, İrem; Tuncay, İbrahim

    2017-01-01

    The purpose of this study is to evaluate the knowledge levels of orthopedic surgeons working in Turkey about the uses and possible risks of fluoroscopy and assess methods for preventing radiation damage. A questionnaire with a total of 12 questions was sent to 1121 orthopedic surgeons working in Turkey. The questionnaire evaluated participants' knowledge about the uses and risks of fluoroscopy and methods for preventing damage. One thousand and twenty-four orthopedic surgeons were found to be suitable for inclusion in the study. The effects of fluoroscopy on patients were not assessed in our study. The data obtained were statistically evaluated. Of the surveyed surgeons, 313 (30%) had used fluoroscopy in over 50% of their operations. The average number of fluoroscopy shots per case was 54.5. A lead apron was the most commonly used (88%) protection from the harmful effects of radiation. Fluoroscopy shots were performed with the help of operating room personnel (86%). A dosimeter was used 5% of the time. According to the survey results, the need for fluoroscopy was very high in orthopedic surgery. However, orthopedic surgeons have inadequate knowledge about the uses and risks of fluoroscopy and methods for preventing damage. Therefore, we believe that training on this topic should be provided to all orthopedic surgeons. © The Authors, published by EDP Sciences, 2017.

  6. Puncture Reduction in Percutaneous Transforaminal Endoscopic Discectomy with HE's Lumbar LOcation (HELLO) System: A Cadaver Study.

    PubMed

    Fan, Guoxin; Guan, Xiaofei; Sun, Qi; Hu, Annan; Zhu, Yanjie; Gu, Guangfei; Zhang, Hailong; He, Shisheng

    2015-01-01

    Percutaneous transforaminal endoscopic discectomy (PTED) usually requires numerous punctures under X-ray fluoroscopy. Repeated puncture will lead to more radiation exposure and reduce the beginners' confidence. This cadaver study aimed to investigate the efficacy of HE's Lumbar Location (HELLO) system in puncture reduction of PTED. Cadaver study. Comparative groups. HELLO system consists of self-made surface locator and puncture locator. One senior surgeon conducted the puncture procedure of PTED on the left side of 20 cadavers at L4/L5 and L5/S1 level with the assistance of HELLO system (Group A). Additionally, the senior surgeon conducted the puncture procedure of PTED on the right side of the cadavers at L4/L5 and L5/S1 level with traditional methods (Group B). On the other hand, an inexperienced surgeon conducted the puncture procedure of PTED on the left side of the cadavers at L4/L5 and L5/S1 level with the assistance of our HELLO system (Group C). At L4/L5 level, there was significant difference in puncture times between Group A and Group B (P<0.001), but no significant difference was observed between Group A and Group C (P = 0.811). Similarly at L5/S1 level, there was significant difference in puncture times between Group A and Group B (P<0.001), but no significant difference was observed between Group A and Group C (P = 0.981). At L4/L5 level, there was significant difference in fluoroscopy time between Group A and Group B (P<0.001), but no significant difference was observed between Group A and Group C (P = 0.290). Similarly at L5/S1 level, there was significant difference in fluoroscopy time between Group A and Group B (P<0.001), but no significant difference was observed between Group A and Group C (P = 0.523). As for radiation exposure, HELLO system reduced 39%-45% radiation dosage when comparing Group A and Group B, but there was no significant difference in radiation exposure between Group A and Group C whatever at L4/L5 level or L5/S1 level (P>0.05). There was no difference in location time between Group A and Group B or Group A and Group C either at L4/L5 level or L5/S1 level (P>0.05). Small-sample preclinical study. HELLO system was effective in reducing puncture times, fluoroscopy time and radiation exposure, as well as the difficulty of learning PTED. (2015-RES-127).

  7. A prototype percutaneous transhepatic cholangiography training simulator with real-time breathing motion.

    PubMed

    Villard, P F; Vidal, F P; Hunt, C; Bello, F; John, N W; Johnson, S; Gould, D A

    2009-11-01

    We present here a simulator for interventional radiology focusing on percutaneous transhepatic cholangiography (PTC). This procedure consists of inserting a needle into the biliary tree using fluoroscopy for guidance. The requirements of the simulator have been driven by a task analysis. The three main components have been identified: the respiration, the real-time X-ray display (fluoroscopy) and the haptic rendering (sense of touch). The framework for modelling the respiratory motion is based on kinematics laws and on the Chainmail algorithm. The fluoroscopic simulation is performed on the graphic card and makes use of the Beer-Lambert law to compute the X-ray attenuation. Finally, the haptic rendering is integrated to the virtual environment and takes into account the soft-tissue reaction force feedback and maintenance of the initial direction of the needle during the insertion. Five training scenarios have been created using patient-specific data. Each of these provides the user with variable breathing behaviour, fluoroscopic display tuneable to any device parameters and needle force feedback. A detailed task analysis has been used to design and build the PTC simulator described in this paper. The simulator includes real-time respiratory motion with two independent parameters (rib kinematics and diaphragm action), on-line fluoroscopy implemented on the Graphics Processing Unit and haptic feedback to feel the soft-tissue behaviour of the organs during the needle insertion.

  8. Transbronchial cryobiopsy in interstitial lung disease: experience in 106 cases – how to do it

    PubMed Central

    Bango-Álvarez, Antonio; Torres-Rivas, Hector; Fernández-Fernández, Luis; Prieto, Amador; Sánchez, Inmaculada; Gil, Maria; Pando-Sandoval, Ana

    2017-01-01

    Transbronchial biopsy using forceps (TBB) is the first diagnostic technique performed on patients with interstitial lung disease (ILD). However, the small size of the samples and the presence of artefacts in the tissue obtained make the yield variable. Our objectives were 1) to attempt to reproduce transbronchial cryobiopsy under the same conditions with which we performed conventional TBB, that is, in the bronchoscopy unit without intubating the patient and without fluoroscopy or general anaesthesia; 2) to describe the method used for its execution; and 3) to analyse the diagnostic yield and its complications. We carried out a prospective study that included 106 patients with clinical and radiological features suggestive of ILD who underwent cryo-transbronchial lung biopsy (cryo-TBB) under moderate sedation without endotracheal intubation, general anaesthesia or use of fluoroscopy. We performed the procedure using two flexible bronchoscopes connected to two video processors, which we alternated until obtaining the number of desired samples. A definitive diagnosis was obtained in 91 patients (86%). As for complications, there were five pneumothoraces (4.7%) and in no case was there severe haemorrhage or exacerbation of the underlying interstitial disease. Cryo-TBB following our method is a minimally invasive, rapid, safe and economic technique that can be performed in a bronchoscopy suite under moderate sedation without the need for intubating the patient or using fluoroscopy and without requiring general anaesthesia. PMID:28344982

  9. Classical fluoroscopy criteria poorly predict right ventricular lead septal positioning by comparison with echocardiography.

    PubMed

    Squara, Fabien; Scarlatti, Didier; Riccini, Philippe; Garret, Gauthier; Moceri, Pamela; Ferrari, Emile

    2018-03-13

    Fluoroscopic criteria have been described for the documentation of septal right ventricular (RV) lead positioning, but their accuracy remains questioned. Consecutive patients undergoing pacemaker or defibrillator implantation were prospectively included. RV lead was positioned using postero-anterior and left anterior oblique 40° incidences, and right anterior oblique 30° to rule out coronary sinus positioning when suspected. RV lead positioning using fluoroscopy was compared to true RV lead positioning as assessed by transthoracic echocardiography (TTE). Precise anatomical localizations were determined with both modalities; then, RV lead positioning was ultimately dichotomized into two simple clinically relevant categories: RV septal or RV free wall. Accuracy of fluoroscopy for RV lead positioning was then assessed by comparison with TTE. We included 100 patients. On TTE, 66/100 had a septal RV lead and 34/100 had a free wall RV lead. Fluoroscopy had moderate agreement with TTE for precise anatomical localization of RV lead (k = 0.53), and poor agreement for septal/free wall localization (k = 0.36). For predicting septal RV lead positioning, classical fluoroscopy criteria had a high sensitivity (95.5%; 63/66 patients having a septal RV lead on TTE were correctly identified by fluoroscopy) but a very low specificity (35.3%; only 12/34 patients having a free wall RV lead on TTE were correctly identified by fluoroscopy). Classical fluoroscopy criteria have a poor accuracy for identifying RV free wall leads, which are most of the time misclassified as septal. This raises important concerns about the efficacy and safety of RV lead positioning using classical fluoroscopy criteria.

  10. Applications for a hybrid operating room in thoracic surgery: from multidisciplinary procedures to ­­image-guided video-assisted thoracoscopic surgery

    PubMed Central

    Terra, Ricardo Mingarini; Andrade, Juliano Ribeiro; Mariani, Alessandro Wasum; Garcia, Rodrigo Gobbo; Succi, Jose Ernesto; Soares, Andrey; Zimmer, Paulo Marcelo

    2016-01-01

    ABSTRACT The concept of a hybrid operating room represents the union of a high-complexity surgical apparatus with state-of-the-art radiological tools (ultrasound, CT, fluoroscopy, or magnetic resonance imaging), in order to perform highly effective, minimally invasive procedures. Although the use of a hybrid operating room is well established in specialties such as neurosurgery and cardiovascular surgery, it has rarely been explored in thoracic surgery. Our objective was to discuss the possible applications of this technology in thoracic surgery, through the reporting of three cases. PMID:27812640

  11. MO-G-18A-01: Radiation Dose Reducing Strategies in CT, Fluoroscopy and Radiography

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Mahesh, M; Gingold, E; Jones, A

    2014-06-15

    Advances in medical x-ray imaging have provided significant benefits to patient care. According to NCRP 160, there are more than 400 million x-ray procedures performed annually in the United States alone that contributes to nearly half of all the radiation exposure to the US population. Similar growth trends in medical x-ray imaging are observed worldwide. Apparent increase in number of medical x-ray imaging procedures, new protocols and the associated radiation dose and risk has drawn considerable attention. This has led to a number of technological innovations such as tube current modulation, iterative reconstruction algorithms, dose alerts, dose displays, flat panelmore » digital detectors, high efficient digital detectors, storage phosphor radiography, variable filters, etc. that are enabling users to acquire medical x-ray images at a much lower radiation dose. Along with these, there are number of radiation dose optimization strategies that users can adapt to effectively lower radiation dose in medical x-ray procedures. The main objectives of this SAM course are to provide information and how to implement the various radiation dose optimization strategies in CT, Fluoroscopy and Radiography. Learning Objectives: To update impact of technological advances on dose optimization in medical imaging. To identify radiation optimization strategies in computed tomography. To describe strategies for configuring fluoroscopic equipment that yields optimal images at reasonable radiation dose. To assess ways to configure digital radiography systems and recommend ways to improve image quality at optimal dose.« less

  12. Respiratory motion compensated overlay of surface models from cardiac MR on interventional x-ray fluoroscopy for guidance of cardiac resynchronization therapy procedures

    NASA Astrophysics Data System (ADS)

    Manzke, R.; Bornstedt, A.; Lutz, A.; Schenderlein, M.; Hombach, V.; Binner, L.; Rasche, V.

    2010-02-01

    Various multi-center trials have shown that cardiac resynchronization therapy (CRT) is an effective procedure for patients with end-stage drug invariable heart failure (HF). Despite the encouraging results of CRT, at least 30% of patients do not respond to the treatment. Detailed knowledge of the cardiac anatomy (coronary venous tree, left ventricle), functional parameters (i.e. ventricular synchronicity) is supposed to improve CRT patient selection and interventional lead placement for reduction of the number of non-responders. As a pre-interventional imaging modality, cardiac magnetic resonance (CMR) imaging has the potential to provide all relevant information. With functional information from CMR optimal implantation target sites may be better identified. Pre-operative CMR could also help to determine whether useful vein target segments are available for lead placement. Fused with X-ray, the mainstay interventional modality, improved interventional guidance for lead-placement could further help to increase procedure outcome. In this contribution, we present novel and practicable methods for a) pre-operative functional and anatomical imaging of relevant cardiac structures to CRT using CMR, b) 2D-3D registration of CMR anatomy and functional meshes with X-ray vein angiograms and c) real-time capable breathing motion compensation for improved fluoroscopy mesh overlay during the intervention based on right ventricular pacer lead tracking. With these methods, enhanced interventional guidance for left ventricular lead placement is provided.

  13. A new method to precisely control the depth of percutaneous screws into the pedicle by counting the rotation number of the screw with low radiation exposure: technical note.

    PubMed

    Li, Xu; Zhang, Feng; Zhang, Wenzhi; Shang, Xifu; Han, Jintao; Liu, Pengfei

    2017-03-01

    Technique note. To report a new method for precisely controlling the depth of percutaneous pedicle screws (PPS)-without radiation exposure to surgeons and less fluoroscopy exposure to patients than with conventional methods. PPS is widely used in minimal invasive spine surgery; the advantages include reduced muscle damage, pain, and hospital stays. However, placement of PPS demands repeated checking with fluoroscopy. Thus, radiation exposure is considerable for both surgeons and patients. The PPS depth was determined by counting rotations of the screws. The distance between screw threads can be measured for particular screws; thus, full rotations of the PPS results in the screw advancing in the pedicle the distance between screw threads. To fully insert screws into the pedicle, the number of full rotations is equal to the number of threads in the PPS. We applied this technique in 58 patients with thoracolumbar fracture. The position and depth of the screws was checked during the operation with the C-arm and after operation by anteroposterior X-ray film or computed tomography. No additional procedures were required to correct the screws; we observed no neurological deficits or malpositioning of the screws. In the screw placement procedure, the radiation exposure for surgeons is zero, and the patient is well protected from extensive radiation exposure. This method of counting rotation of screws is a safe way to precisely determine the depth of PPS in the placement procedure. IV.

  14. Technical Note: Thoracic duct embolization for treatment of chylothorax: A novel guidance technique for puncture using combined MRI and fluoroscopy.

    PubMed

    Praveen, Alampath; Sreekumar, Karumathil Pullara; Nazar, Puthukudiyil Kader; Moorthy, Srikanth

    2012-04-01

    Thoracic duct embolization (TDE) is an established radiological interventional procedure for thoracic duct injuries. Traditionally, it is done under fluoroscopic guidance after opacifying the thoracic duct with bipedal lymphangiography. We describe our experience in usinga heavily T2W sequence for guiding thoracic duct puncture and direct injection of glue through the puncture needle without cannulating the duct.

  15. Efficacy comparison between cryoablation and radiofrequency ablation for patients with cavotricuspid valve isthmus dependent atrial flutter: a meta-analysis

    NASA Astrophysics Data System (ADS)

    Chen, Yi-He; Lin, Hui; Xie, Cheng-Long; Zhang, Xiao-Ting; Li, Yi-Gang

    2015-06-01

    We perform this meta-analysis to compare the efficacy and safety of cryoablation versus radiofrequency ablation for patients with cavotricuspid valve isthmus dependent atrial flutter. By searching EMBASE, MEDLINE, PubMed and Cochrane electronic databases from March 1986 to September 2014, 7 randomized clinical trials were included. Acute (risk ratio[RR]: 0.93; P = 0.14) and long-term (RR: 0.94; P = 0.08) success rate were slightly lower in cryoablation group than in radiofrequency ablation group, but the difference was not statistically significant. Additionally, the fluoroscopy time was nonsignificantly reduced (weighted mean difference[WMD]: -2.83 P = 0.29), whereas procedure time was significantly longer (WMD: 25.95; P = 0.01) in cryoablation group compared with radiofrequency ablation group. Furthermore, Pain perception during the catheter ablation was substantially less in cryoabaltion group than in radiofrequency ablation group (standardized mean difference[SMD]: -2.36 P < 0.00001). Thus, our meta-analysis demonstrated that cryoablation and radiofrequency ablation produce comparable acute and long-term success rate for patients with cavotricuspid valve isthmus dependent atrial flutter. Meanwhile, cryoablation ablation tends to reduce the fluoroscopy time and significantly reduce pain perception in cost of significantly prolonged procedure time.

  16. Efficacy comparison between cryoablation and radiofrequency ablation for patients with cavotricuspid valve isthmus dependent atrial flutter: a meta-analysis

    PubMed Central

    Chen, Yi-He; Lin, Hui; Xie, Cheng-Long; Zhang, Xiao-Ting; Li, Yi-Gang

    2015-01-01

    We perform this meta-analysis to compare the efficacy and safety of cryoablation versus radiofrequency ablation for patients with cavotricuspid valve isthmus dependent atrial flutter. By searching EMBASE, MEDLINE, PubMed and Cochrane electronic databases from March 1986 to September 2014, 7 randomized clinical trials were included. Acute (risk ratio[RR]: 0.93; P = 0.14) and long-term (RR: 0.94; P = 0.08) success rate were slightly lower in cryoablation group than in radiofrequency ablation group, but the difference was not statistically significant. Additionally, the fluoroscopy time was nonsignificantly reduced (weighted mean difference[WMD]: −2.83; P = 0.29), whereas procedure time was significantly longer (WMD: 25.95; P = 0.01) in cryoablation group compared with radiofrequency ablation group. Furthermore, Pain perception during the catheter ablation was substantially less in cryoabaltion group than in radiofrequency ablation group (standardized mean difference[SMD]: −2.36; P < 0.00001). Thus, our meta-analysis demonstrated that cryoablation and radiofrequency ablation produce comparable acute and long-term success rate for patients with cavotricuspid valve isthmus dependent atrial flutter. Meanwhile, cryoablation ablation tends to reduce the fluoroscopy time and significantly reduce pain perception in cost of significantly prolonged procedure time. PMID:26039980

  17. On-line data collection platform for national dose surveys in diagnostic and interventional radiology.

    PubMed

    Vassileva, J; Simeonov, F; Avramova-Cholakova, S

    2015-07-01

    According to the Bulgarian regulation for radiation protection at medical exposure, the National Centre of Radiobiology and Radiation Protection (NCRRP) is responsible for performing national dose surveys in diagnostic and interventional radiology and nuclear medicine and for establishing of national diagnostic reference levels (DRLs). The next national dose survey is under preparation to be performed in the period of 2015-16, with the aim to cover conventional radiography, mammography, conventional fluoroscopy, interventional and fluoroscopy guided procedures and CT. It will be performed electronically using centralised on-line data collection platform established by the NCRRP. The aim is to increase the response rate and to improve the accuracy by reducing human errors. The concept of the on-line dose data collection platform is presented. Radiological facilities are provided with a tool to determine local typical patient doses, and the NCRRP to establish national DRLs. Future work will include automatic retrieval of dose data from hospital picture archival and communicating system. The on-line data collection platform is expected to facilitate the process of dose audit and optimisation of radiological procedures in Bulgarian hospitals. © The Author 2015. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  18. The Effect of Realtime Monitoring on Dose Exposure to Staff Within an Interventional Radiology Setting

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Baumann, Frederic, E-mail: fredericbaumann@hotmail.com; Katzen, Barry T.; Carelsen, Bart

    PurposeThe purpose of this study is to evaluate a new device providing real-time monitoring on radiation exposure during fluoroscopy procedures intending to reduce radiation in an interventional radiology setting.Materials and MethodsIn one interventional suite, a new system providing a real-time radiation dose display and five individual wireless dosimeters were installed. The five dosimeters were worn by the attending, fellow, nurse, technician, and anesthesiologist for every procedure taking place in that suite. During the first 6-week interval the dose display was off (closed phase) and activated thereafter, for a 6-week learning phase (learning phase) and a 10-week open phase (open phase).more » During these phases, the staff dose and the individual dose for each procedure were recorded from the wireless dosimeter and correlated with the fluoroscopy time. Further subanalysis for dose exposure included diagnostic versus interventional as well as short (<10 min) versus long (>10 min) procedures.ResultsA total of 252 procedures were performed (n = 88 closed phase, n = 50 learning phase, n = 114 open phase). The overall mean staff dose per fluoroscopic minute was 42.79 versus 19.81 µSv/min (p < 0.05) comparing the closed and open phase. Thereby, anesthesiologists were the only individuals attaining a significant dose reduction during open phase 16.9 versus 8.86 µSv/min (p < 0.05). Furthermore, a significant reduction of total staff dose was observed for short 51 % and interventional procedures 45 % (p < 0.05, for both).ConclusionA real-time qualitative display of radiation exposure may reduce team radiation dose. The process may take a few weeks during the learning phase but appears sustained, thereafter.« less

  19. A Novel Inherently Radiopaque Bead for Transarterial Embolization to Treat Liver Cancer - A Pre-clinical Study.

    PubMed

    Duran, Rafael; Sharma, Karun; Dreher, Matthew R; Ashrafi, Koorosh; Mirpour, Sahar; Lin, MingDe; Schernthaner, Ruediger E; Schlachter, Todd R; Tacher, Vania; Lewis, Andrew L; Willis, Sean; den Hartog, Mark; Radaelli, Alessandro; Negussie, Ayele H; Wood, Bradford J; Geschwind, Jean-François H

    2016-01-01

    Embolotherapy using microshperes is currently performed with soluble contrast to aid in visualization. However, administered payload visibility dimishes soon after delivery due to soluble contrast washout, leaving the radiolucent bead's location unknown. The objective of our study was to characterize inherently radiopaque beads (RO Beads) in terms of physicomechanical properties, deliverability and imaging visibility in a rabbit VX2 liver tumor model. RO Beads, which are based on LC Bead® platform, were compared to LC Bead. Bead size (light microscopy), equilibrium water content (EWC), density, X-ray attenuation and iodine distribution (micro-CT), suspension (settling times), deliverability and in vitro penetration were investigated. Fifteen rabbits were embolized with either LC Bead or RO Beads + soluble contrast (iodixanol-320), or RO Beads+dextrose. Appearance was evaluated with fluoroscopy, X-ray single shot, cone-beam CT (CBCT). Both bead types had a similar size distribution. RO Beads had lower EWC (60-72%) and higher density (1.21-1.36 g/cc) with a homogeneous iodine distribution within the bead's interior. RO Beads suspension time was shorter than LC Bead, with durable suspension (>5 min) in 100% iodixanol. RO Beads ≤300 µm were deliverable through a 2.3-Fr microcatheter. Both bead types showed similar penetration. Soluble contrast could identify target and non-target embolization on fluoroscopy during administration. However, the imaging appearance vanished quickly for LC Bead as contrast washed-out. RO Beads+contrast significantly increased visibility on X-ray single shot compared to LC Bead+contrast in target and non-target arteries (P=0.0043). Similarly, RO beads demonstrated better visibility on CBCT in target arteries (P=0.0238) with a trend in non-target arteries (P=0.0519). RO Beads+dextrose were not sufficiently visible to monitor embolization using fluoroscopy. RO Beads provide better conspicuity to determine target and non-target embolization compared to LC Bead which may improve intra-procedural monitoring and post-procedural evaluation of transarterial embolization.

  20. Image Registration of Cone-Beam Computer Tomography and Preprocedural Computer Tomography Aids in Localization of Adrenal Veins and Decreasing Radiation Dose in Adrenal Vein Sampling

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Busser, Wendy M. H., E-mail: wendy.busser@radboudumc.nl; Arntz, Mark J.; Jenniskens, Sjoerd F. M.

    2015-08-15

    PurposeWe assessed whether image registration of cone-beam computed tomography (CT) (CBCT) and contrast-enhanced CT (CE-CT) images indicating the locations of the adrenal veins can aid in increasing the success rate of first-attempts adrenal vein sampling (AVS) and therefore decreasing patient radiation dose.Materials and Methods CBCT scans were acquired in the interventional suite (Philips Allura Xper FD20) and rigidly registered to the vertebra in previously acquired CE-CT. Adrenal vein locations were marked on the CT image and superimposed with live fluoroscopy and digital-subtraction angiography (DSA) to guide the AVS. Seventeen first attempts at AVS were performed with image registration and retrospectivelymore » compared with 15 first attempts without image registration performed earlier by the same 2 interventional radiologists. First-attempt AVS was considered successful when both adrenal vein samples showed representative cortisol levels. Sampling time, dose-area product (DAP), number of DSA runs, fluoroscopy time, and skin dose were recorded.ResultsWithout image registration, the first attempt at sampling was successful in 8 of 15 procedures indicating a success rate of 53.3 %. This increased to 76.5 % (13 of 17) by adding CBCT and CE-CT image registration to AVS procedures (p = 0.266). DAP values (p = 0.001) and DSA runs (p = 0.026) decreased significantly by adding image registration guidance. Sampling and fluoroscopy times and skin dose showed no significant changes.ConclusionGuidance based on registration of CBCT and previously acquired diagnostic CE-CT can aid in enhancing localization of the adrenal veins thereby increasing the success rate of first-attempt AVS with a significant decrease in the number of used DSA runs and, consequently, radiation dose required.« less

  1. SU-E-I-58: Experiences in Setting Up An Online Fluoroscopy Tracking System in a Large Healthcare System

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fisher, R; Wunderle, K; Lingenfelter, M

    Purpose: Transitioning from a paper based to an online system for tracking fluoroscopic case information required by state regulation and to conform to NCRP patient dose tracking suggestions. Methods: State regulations require documentation of operator, equipment, and some metric of tube output for fluoroscopy exams. This information was previously collected in paper logs, which was cumbersome and inefficient for the large number of fluoroscopic units across multiple locations within the system. The “tech notes” feature within Siemens’ Syngo workflow RIS was utilized to create an entry form for technologists to input case information, which was sent to a third partymore » vendor for archiving and display though an online web based portal. Results: Over 55k cases were logged in the first year of implementation, with approximately 6,500 cases per month once fully online. A system was built for area managers to oversee and correct data, which has increased the accuracy of inputted values. A high-dose report was built to automatically send notifications when patients exceed trigger levels. In addition to meeting regulatory requirements, the new system allows for larger scale QC in fluoroscopic cases by allowing comparison of data from specific procedures, locations, equipment, and operators so that instances that fall outside of reference levels can be identified for further evaluation. The system has also drastically improved identification of operators without documented equipment specific training. Conclusion: The transition to online fluoroscopy logs has improved efficiency in meeting state regulatory requirements as well as allowed for identification of particular procedures, equipment, and operators in need of additional attention in order to optimize patient and personnel doses, while high dose alerts improve patient care and follow up. Future efforts are focused on incorporating case information from outside of radiology, as well as on automating processes for increased efficiencies.« less

  2. Characterization of MOSFET Dosimeter Angular Response Using a Spherical Phantom for Fluoroscopic Dosimetry.

    PubMed

    Wang, Chu; Hill, Kevin; Yoshizumi, Terry

    2016-01-01

    Metal-Oxide Semiconductor Field-Effect Transistor (MOSFET) dosimeters, placed in anthropomorphic phantoms, are a standard method for organ dosimetry in medical x-ray imaging applications. However, many x-ray applications, particularly fluoroscopy procedures, use variable projection angles. During dosimetry, the MOSFET detector active area may not always be perpendicular to the x-ray beam. The goal of this study was to characterize the dosimeter's angular response in the fluoroscopic irradiation involved in pediatric cardiac catheterization procedures, during which a considerable amount of fluoroscopic x-ray irradiation is often applied from various projection angles. A biological x-ray irradiator was used to simulate the beam quality of a biplane fluoroscopy imaging system. A custom-designed acrylic spherical scatter phantom was fabricated to measure dosimeter response (in mV) in two rotational axes, axial (ψ) and normal-to-axial (θ), in 30° increments, as well as four common oblique angles used in cardiac catheterization: a) 90° Left Anterior Oblique (LAO); b) 70° LAO/ 20° Cranial; c) 20° LAO/ 15° Cranial; and d) 30° Right Anterior Oblique (RAO). All results were normalized to the angle where the dosimeter epoxy is perpendicular to the beam or the Posterior-Anterior projection angle in the clinical setup. The relative response in the axial rotation was isotropic (within ± 10% deviation); that in the normal-to-axial rotation was isotropic in all angles except the ψ = 270° angle, where the relative response was 83 ± 9%. No significant deviation in detector response was observed in the four common oblique angles, with their relative responses being: a) 102 ± 3%; b) 90 ± 3%; c) 92 ± 3%; and d) 95 ± 3%, respectively. These angular correction factors will be used in future dosimetry studies for fluoroscopy. The spherical phantom may be useful for other applications, as it allows the measurement of dosimeter response in virtually all angles in the 3-dimensional spherical coordinates.

  3. Flexible Ureterorenoscopy versus Mini-Percutaneous Nephrolithotomy for the Treatment of Renal Stones.

    PubMed

    Ergin, Giray; Kirac, Mustafa; Kopru, Burak; Ebiloglu, Turgay; Biri, Hasan

    2018-04-22

    To compare the pain status and stone free rates of flexible ureterorenoscopy (F-URS) versus mini-percutaneousnephrolithotomy (mini-PNL) for the treatment of 1-to 2-cm renal stones. This study was retrospectively designed with match paired method. Between January 2013 and December 2016, 387 patients underwent stone surgery for renal stones, 45 patients underwent FURS and 45 patients underwent mini-PNL. 90 patients were divided into two groups according to the surgical procedures. Group 1 patients underwent F-URS, and Group 2 patients underwent mini-PNL. During the intraoperative andpostoperative periods, pain management for all patients was standardized. Pain scores were determined using a visual analogue scale (VAS) completed at 2, 6, 12 and 24 hours postoperatively. The stone free status, hemoglobin levels, fluoroscopy time (FT), operation time (OT), hospitalization time (HT), return to work time (RWT), and complications were noted for each patient. Of all patients, the mean age was 41.1 ± 12.1 years and the mean stone size was 13.9 ± 2.9 mm. The VAS scores were significantly higher in the mini-PNL group at 2, 6, 12 and 24 hours (P < .05). The stone-free status and complication rates were similar between the two groups (P > .05); however, the hemoglobin decreases and the fluoroscopy, operation, hospitalization and return to work times were higher in the mini-PNL group than in the F-URS group (P < .05). F-URS is less painful than mini-PNL for the treatment of 1- to 2-cm renal stones. However, the stone free rate is similar between the two procedures while mini-PNL is superior in terms of fluoroscopy, operation, hospitalization and return to work duration. We think that F-URS is more comfortable and less painful than mini-PNL and achieves a similar stone free rate for the treatment of 1- to 2-cm renal stones.

  4. Approach of forensic medicine to gossypiboma.

    PubMed

    Karakaya, M Arif; Koç, Okay; Ekiz, Feza; Ağaçhan, A Feran

    2015-01-01

    The aim of this study was to assess the risk factors and preventive measures for gossypibomas and their medico-legal implications in forensic medicine in the Turkish legal system. This study involved a retrospective analysis of the records of 39 patients with gossypiboma. Records were available from the Istanbul Forensic Medicine Institution and were surveyed for faulty treatment between 2008 and 2012. Parameters such as distribution of the cases according to specializations, elective and emergency procedures, surgical procedures, radio-opaque sponge and fluoroscopy availability, routine sponge and instrument counting, number of nurses for counting, and control of the operative field by a second surgeon were investigated. All cases were evaluated by the Istanbul Forensic Medicine Institute 3rd Expertise Committee. This committee comprised of specialists from the departments of forensic medicine, orthopedics and traumatology, general surgery, neurology, internal medicine, pediatrics, chest disease, and infectious diseases. All cases were considered as poor medical practice (malpractice) and surgeons were found to be responsible. In 16 of these 39 cases (41%) emergency procedures were performed. No unexpected event was reported in any procedure. In 16 cases (41%), sponge count was performed and was reported to be complete. Operation notes were available in 16 (41%) cases. Control of the operative field was performed by 1 surgeon, and sponge and instrument count was performed by 1 scrub nurse. Radio-opaque sponge and fluoroscopy were available in 9 (23%) centers in these cases. Gossypiboma can be prevented not only with surgeons' care but also with adequate support of medical device and material. However, it is considered as a poor medical practice. Presence of only 1 general surgeon in the expertise committee and ignorance of the working conditions by the surgeons should be questioned.

  5. Plug-Assisted Retrograde Transvenous Obliteration for the Treatment of Gastric Varices: The Role of Intra-Procedural Cone-Beam Computed Tomography

    PubMed Central

    Ko, Gi-Young; Kwon, Young Baek; Yoon, Hyun-Ki; Sung, Kyu-Bo

    2018-01-01

    Objective To investigate the technical and clinical outcomes of plug-assisted retrograde transvenous obliteration (PARTO) for the treatment of gastric varices (GV) and to evaluate the role of intra-procedural cone-beam computed tomography (CBCT) performed during PARTO to confirm its technical success. Materials and Methods From January 2016 to December 2016, 17 patients with GV who had undergone PARTO were retrospectively evaluated. When the proximal part of the afferent vein was identified on a fluoroscopy, non-contrast CBCT images were obtained. In patients with incomplete embolization of GV, an additional injection of gelatin sponges was performed. Follow-up data from contrast-enhanced CT and upper intestinal endoscopy, as well as clinical and laboratory data were collected. Results Plug-assisted retrograde transvenous obliteration procedures were technically successful in all 17 patients. Complete embolization of GV was detected on CBCT images in 15 patients; whereas, incomplete embolization was detected in two. Complete embolization of GV was then achieved after an additional injection of gelatin sponges in these two patients as demonstrated on the 2nd CBCT image. The mean follow-up period after PARTO was 193 days (range, 73–383 days). A follow-up CT obtained 2–4 months after PARTO demonstrated marked shrinkage or complete obliteration of GV and portosystemic shunts in all 17 patients. There were no cases of variceal bleeding during the follow-up. Conclusion Plug-assisted retrograde transvenous obliteration is technically and clinically effective for the treatment of GV. In addition, intra-procedural CBCT can be an adjunct tool to fluoroscopy, because it can provide an immediate and accurate evaluation of the technical success of PARTO. PMID:29520179

  6. Dual-Axis Rotational Angiography is Safe and Feasible to Detect Coronary Allograft Vasculopathy in Pediatric Heart Transplant Patients: A Single-Center Experience.

    PubMed

    Rios, Rodrigo; Loomba, Rohit S; Foerster, Susan R; Pelech, Andrew N; Gudausky, Todd M

    2016-04-01

    Coronary allograft vasculopathy (CAV) is the leading cause of graft failure in pediatric heart transplant recipients, also adding to mortality in this patient population. Coronary angiography is routinely performed to screen for CAV, with conventional single-plane or bi-plane angiography being utilized. Dual-axis rotational coronary angiography (RA) has been described, mostly in the adult population, and may offer reduction in radiation dose and contrast volume. Experience with this in the pediatric population is limited. This study describes a single-institution experience with RA for screening for CAV in pediatric patients. The catheterization database at our institution was used to identify pediatric heart transplant recipients having undergone RA to screen for CAV. Procedural data including radiation dose, fluoroscopy time, contrast volume, and procedure time were collected for each catheterization. The number of instances in which RA was not successful, ECG changes were present, and CAV was detected were also collected for each catheterization. A total of 97 patients underwent 345 catheterizations utilizing RA. Median radiation dose-area product per kilogram was found to be 341.7 (mGy cm(2)/kg), total air kerma was 126.8 (mGy), procedure time was 69 min, fluoroscopy time was 9.9 min, and contrast volume was 13 ml. A total of 17 (2 %) coronary artery injections out of 690 could not be successfully imaged using RA. A total of 14 patients had CAV noted at any point, 10 of whom had progressive CAV. Electrocardiographic changes were documented in a total of 10 (3 %) RA catheterizations. Procedural characteristics did not differ between serial catheterizations. RA is safe and feasible for CAV screening in pediatric heart transplant recipients while offering coronary imaging in multiple planes compared to conventional angiography.

  7. Remote magnetic versus manual catheter navigation for circumferential pulmonary vein ablation in patients with atrial fibrillation.

    PubMed

    Lüthje, Lars; Vollmann, Dirk; Seegers, Joachim; Dorenkamp, Marc; Sohns, Christian; Hasenfuss, Gerd; Zabel, Markus

    2011-11-01

    Only limited data exist on the clinical utility of remote magnetic navigation (RMN) for pulmonary vein (PV) ablation. Aim of this prospective study was to evaluate the safety and efficacy of RMN for PV isolation as compared to the manual (CON) approach. A total of 161 consecutive patients undergoing circumferential PV isolation were included. Open-irrigated 3.5 mm ablation catheters under the guidance of a mapping system were used. The catheter was navigated with the Stereotaxis Niobe II system in the RMN group (n = 107) and guided manually in the CON group (n = 54). Electrical isolation of all PVs was achieved in 90% of the patients in the RMN group and in 87% in the CON group (p = 0.6). All subjects were followed every 3 months by 7d Holter-ECG. At 12 months of follow-up, 53.5% (RMN) and 55.5% (CON) of the patients were free of any left atrial tachycardia/atrial fibrillation (AF) episode (p = 0.57). Free of symptomatic AF recurrence were 66.3% (RMN) and 62.1% (CON) of the subjects (p = 0.80). Use of RMN was associated with longer procedure duration (p < 0.0001), ablation times (p < 0.0001), and RF current application duration (p < 0.05). In contrast, fluoroscopy time was lower in the RMN group (p < 0.0001). Major complications occurred in 6 of 161 procedures (3.7%), with no significant difference between groups (p = 0.75). RMN-guided PV ablation provides comparable acute and long-term success rates as compared to manual navigation. Procedural complication rates are similar. The use of RMN is associated with markedly reduced fluoroscopy time, but prolonged ablation and procedure duration.

  8. Atrial fibrillation ablation: "perpetual motion" of open irrigated tip catheters at 50 W is safe and improves outcomes.

    PubMed

    Winkle, Roger A; Mead, R Hardwin; Engel, Gregory; Patrawala, Rob A

    2011-05-01

    Point-by-point use of open irrigated tip catheters (OITCs) at 50 W increases atrial fibrillation (AF) ablation cure rates but also increases complications. We determined if constantly moving the OITC (perpetual motion) when using 50 W increases ablation cure rates without increasing complications. We evaluated procedural data, complications, and individual procedure cure rates (IPCRs) for AF ablation using closed tip catheters (CTC) versus OITC at 40, 45, and 50 W in 1,122 ablations. We used "perpetual motion" to move the OITC at 50 W every 3-10 seconds. The OITC showed higher IPCR than CTC at 45 W (P = 0.012) and 50 W (P < 0.0005). For the OITC, IPCR increased from 44.6% to 60.7% as power increased from 40 to 50 W (P = 0.008). The OITC appeared superior to the CTC for all types of AF. For paroxysmal AF, increasing OITC power from 40 to 50 W provided no increase in IPCR (70.6% vs 71.2%, P = 0.827). For persistent AF, increasing power from 40 to 50 W increased IPCR from 34.5% to 59.5% (P = 0.001). Complications were similar for the CTC and the OITC at any power. The OITC at 50 W had shorter procedure, left atrial, and fluoroscopy times (P < 0.0005). Increasing OITC power from 40 to 50 W increases IPCR with no increase in complications as long as the 50 W setting is done using "perpetual motion." The OITC 50 W power setting results in shorter procedure and fluoroscopy times and should be considered for AF ablations. ©2010, The Authors. Journal compilation ©2010 Wiley Periodicals, Inc.

  9. Unenhanced Cone Beam Computed Tomography and Fusion Imaging in Direct Percutaneous Sac Injection for Treatment of Type II Endoleak: Technical Note

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Carrafiello, Gianpaolo, E-mail: gcarraf@gmail.com; Ierardi, Anna Maria; Radaelli, Alessandro

    AimTo evaluate safety, feasibility, technical success, and clinical success of direct percutaneous sac injection (DPSI) for the treatment of type II endoleaks (T2EL) using anatomical landmarks on cone beam computed tomography (CBCT) and fusion imaging (FI).Materials and MethodsEight patients with T2EL were treated with DPSI using CBCT as imaging guidance. Anatomical landmarks on unenhanced CBCT were used for referencing T2EL location in the first five patients, while FI between unenhanced CBCT and pre-procedural computed tomography angiography (CTA) was used in the remaining three patients. Embolization was performed with thrombin, glue, and ethylene–vinyl alcohol copolymer. Technical and clinical success, iodinated contrastmore » utilization, procedural time, fluoroscopy time, and mean radiation dose were registered.ResultsDPSI was technically successful in all patients: the needle was correctly positioned at the first attempt in six patients, while in two of the first five patients the needle was repositioned once. Neither minor nor major complications were registered. Average procedural time was 45 min and the average administered iodinated contrast was 13 ml. Mean radiation dose of the procedure was 60.43 Gy cm{sup 2} and mean fluoroscopy time was 18 min. Clinical success was achieved in all patients (mean follow-up of 36 months): no sign of T2EL was reported in seven patients until last CT follow-up, while it persisted in one patient with stability of sac diameter.ConclusionsDPSI using unenhanced CBCT and FI is feasible and provides the interventional radiologist with an accurate and safe alternative to endovascular treatment with limited iodinated contrast utilization.« less

  10. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fetterly, K; Mathew, V

    Purpose: Transcatheter aortic valve replacement (TAVR) procedures provide a method to implant a prosthetic aortic valve via a minimallyinvasive, catheter-based procedure. TAVR procedures require use of interventional fluoroscopy c-arm projection angles which are aligned with the aortic valve plane to minimize prosthetic valve positioning error due to x-ray imaging parallax. The purpose of this work is to calculate the continuous range of interventional fluoroscopy c-arm projection angles which are aligned with the aortic valve plane from a single planar image of a valvuloplasty balloon inflated across the aortic valve. Methods: Computational methods to measure the 3D angular orientation of themore » aortic valve were developed. Required inputs include a planar x-ray image of a known valvuloplasty balloon inflated across the aortic valve and specifications of x-ray imaging geometry from the DICOM header of the image. A-priori knowledge of the species-specific typical range of aortic orientation is required to specify the sign of the angle of the long axis of the balloon with respect to the x-ray beam. The methods were validated ex-vivo and in a live pig. Results: Ex-vivo experiments demonstrated that the angular orientation of a stationary inflated valvuloplasty balloon can be measured with precision less than 1 degree. In-vivo pig experiments demonstrated that cardiac motion contributed to measurement variability, with precision less than 3 degrees. Error in specification of x-ray geometry directly influences measurement accuracy. Conclusion: This work demonstrates that the 3D angular orientation of the aortic valve can be calculated precisely from a planar image of a valvuloplasty balloon inflated across the aortic valve and known x-ray geometry. This method could be used to determine appropriate c-arm angular projections during TAVR procedures to minimize x-ray imaging parallax and thereby minimize prosthetic valve positioning errors.« less

  11. Technical Note: Thoracic duct embolization for treatment of chylothorax: A novel guidance technique for puncture using combined MRI and fluoroscopy

    PubMed Central

    Praveen, Alampath; Sreekumar, Karumathil Pullara; Nazar, Puthukudiyil Kader; Moorthy, Srikanth

    2012-01-01

    Thoracic duct embolization (TDE) is an established radiological interventional procedure for thoracic duct injuries. Traditionally, it is done under fluoroscopic guidance after opacifying the thoracic duct with bipedal lymphangiography. We describe our experience in usinga heavily T2W sequence for guiding thoracic duct puncture and direct injection of glue through the puncture needle without cannulating the duct. PMID:23162248

  12. Organ dose assessment in pediatric fluoroscopy and CT via a tomographic computational phantom of the newborn patient

    NASA Astrophysics Data System (ADS)

    Staton, Robert J.

    Of the various types of imaging modalities used in pediatric radiology, fluoroscopy and computed tomography (CT) have the highest associated radiation dose. While these examinations are commonly used for pediatric patients, little data exists on the magnitude of the organ and effective dose values for these procedures. Calculation of these dose values is necessary because of children's increased sensitivity to radiation and their long life expectancy for which to express radiation's latent effects. In this study, a newborn tomographic phantom has been implemented in a radiation transport code to evaluate organ and effective doses for newborn patients in commonly performed fluoroscopy and CT examinations. Organ doses were evaluated for voiding cystourethrogram (VCUG) fluoroscopy studies of infant patients. Time-sequence analysis was performed for videotaped VCUG studies of five different patients. Organ dose values were then estimated for each patient through Monte Carlo (MC) simulations. The effective dose values of the VCUG examination for five patients ranged from 0.6 mSv to 3.2 mSv, with a mean of 1.8 +/- 0.9 mSv. Organ doses were also assessed for infant upper gastrointestinal (UGI) fluoroscopy exams. The effective dose values of the UGI examinations for five patients ranged from 1.05 mSv to 5.92 mSv, with a mean of 2.90 +/- 1.97 mSv. MC simulations of helical multislice CT (MSCT) exams were also completed using, the newborn tomographic phantom and a stylized newborn phantom. The helical path of the source, beam shaping filter, beam profile, patient table, were all included in the MC simulations of the helical MSCT scanner. Organ doses and effective doses and their dependence on scan parameters were evaluated for newborn patients. For all CT scans, the effective dose was found to range approximately 1-13 mSv, with the largest values occurring for CAP scans. Tube current modulation strategies to reduce patient dose were also evaluated for newborn patients. Overall, utilization of the newborn tomographic phantom in MC simulations has shown the need for and usefulness of pediatric tomographic phantoms. The newborn tomographic model has shown more versatility and realistic anatomical modeling when compared to the existing stylized newborn phantom. This work has provided important organ dose data for infant patients in common examinations in pediatric radiology.

  13. Depth-resolved registration of transesophageal echo to x-ray fluoroscopy using an inverse geometry fluoroscopy system

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hatt, Charles R.; Tomkowiak, Michael T.; Dunkerley, David A. P.

    2015-12-15

    Purpose: Image registration between standard x-ray fluoroscopy and transesophageal echocardiography (TEE) has recently been proposed. Scanning-beam digital x-ray (SBDX) is an inverse geometry fluoroscopy system designed for cardiac procedures. This study presents a method for 3D registration of SBDX and TEE images based on the tomosynthesis and 3D tracking capabilities of SBDX. Methods: The registration algorithm utilizes the stack of tomosynthetic planes produced by the SBDX system to estimate the physical 3D coordinates of salient key-points on the TEE probe. The key-points are used to arrive at an initial estimate of the probe pose, which is then refined using amore » 2D/3D registration method adapted for inverse geometry fluoroscopy. A phantom study was conducted to evaluate probe pose estimation accuracy relative to the ground truth, as defined by a set of coregistered fiducial markers. This experiment was conducted with varying probe poses and levels of signal difference-to-noise ratio (SDNR). Additional phantom and in vivo studies were performed to evaluate the correspondence of catheter tip positions in TEE and x-ray images following registration of the two modalities. Results: Target registration error (TRE) was used to characterize both pose estimation and registration accuracy. In the study of pose estimation accuracy, successful pose estimates (3D TRE < 5.0 mm) were obtained in 97% of cases when the SDNR was 5.9 or higher in seven out of eight poses. Under these conditions, 3D TRE was 2.32 ± 1.88 mm, and 2D (projection) TRE was 1.61 ± 1.36 mm. Probe localization error along the source-detector axis was 0.87 ± 1.31 mm. For the in vivo experiments, mean 3D TRE ranged from 2.6 to 4.6 mm and mean 2D TRE ranged from 1.1 to 1.6 mm. Anatomy extracted from the echo images appeared well aligned when projected onto the SBDX images. Conclusions: Full 6 DOF image registration between SBDX and TEE is feasible and accurate to within 5 mm. Future studies will focus on real-time implementation and application-specific analysis.« less

  14. How Does Patient Radiation Exposure Compare With Low-dose O-arm Versus Fluoroscopy for Pedicle Screw Placement in Idiopathic Scoliosis?

    PubMed

    Su, Alvin W; McIntosh, Amy L; Schueler, Beth A; Milbrandt, Todd A; Winkler, Jennifer A; Stans, Anthony A; Larson, A Noelle

    Intraoperative C-arm fluoroscopy and low-dose O-arm are both reasonable means to assist in screw placement for idiopathic scoliosis surgery. Both using pediatric low-dose O-arm settings and minimizing the number of radiographs during C-arm fluoroscopy guidance decrease patient radiation exposure and its deleterious biological effect that may be associated with cancer risk. We hypothesized that the radiation dose for C-arm-guided fluoroscopy is no less than low-dose O-arm scanning for placement of pedicle screws. A multicenter matched-control cohort study of 28 patients in total was conducted. Fourteen patients who underwent O-arm-guided pedicle screw insertion for spinal fusion surgery in 1 institution were matched to another 14 patients who underwent C-arm fluoroscopy guidance in the other institution in terms of the age of surgery, body weight, and number of imaged spine levels. The total effective dose was compared. A low-dose pediatric protocol was used for all O-arm scans with an effective dose of 0.65 mSv per scan. The effective dose of C-arm fluoroscopy was determined using anthropomorphic phantoms that represented the thoracic and lumbar spine in anteroposterior and lateral views, respectively. The clinical outcome and complications of all patients were documented. The mean total effective dose for the O-arm group was approximately 4 times higher than that of the C-arm group (P<0.0001). The effective dose for the C-arm patients had high variability based on fluoroscopy time and did not correlate with the number of imaged spine levels or body weight. The effective dose of 1 low-dose pediatric O-arm scan approximated 85 seconds of the C-arm fluoroscopy time. All patients had satisfactory clinical outcomes without major complications that required returning to the operating room. Radiation exposure required for O-arm scans can be higher than that required for C-arm fluoroscopy, but it depends on fluoroscopy time. Inclusion of more medical centers and surgeons will better account for the variability of C-arm dose due to distinct patient characteristics, surgeon's preference, and individual institution's protocol. Level III-case-control study.

  15. MO-DE-BRA-04: Hands-On Fluoroscopy Safety Training with Real-Time Patient and Staff Dosimetry

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Vanderhoek, M; Bevins, N

    Purpose: Fluoroscopically guided interventions (FGI) are routinely performed across many different hospital departments. However, many involved staff members have minimal training regarding safe and optimal use of fluoroscopy systems. We developed and taught a hands-on fluoroscopy safety class incorporating real-time patient and staff dosimetry in order to promote safer and more optimal use of fluoroscopy during FGI. Methods: The hands-on fluoroscopy safety class is taught in an FGI suite, unique to each department. A patient equivalent phantom is set on the patient table with an ion chamber positioned at the x-ray beam entrance to the phantom. This provides a surrogatemore » measure of patient entrance dose. Multiple solid state dosimeters (RaySafe i2 dosimetry systemTM) are deployed at different distances from the phantom (0.1, 1, 3 meters), which provide surrogate measures of staff dose. Instructors direct participating clinical staff to operate the fluoroscopy system as they view live fluoroscopic images, patient entrance dose, and staff doses in real-time. During class, instructors work with clinical staff to investigate how patient entrance dose, staff doses, and image quality are affected by different parameters, including pulse rate, magnification, collimation, beam angulation, imaging mode, system geometry, distance, and shielding. Results: Real-time dose visualization enables clinical staff to directly see and learn how to optimize their use of their own fluoroscopy system to minimize patient and staff dose, yet maintain sufficient image quality for FGI. As a direct result of the class, multiple hospital departments have implemented changes to their imaging protocols, including reduction of the default fluoroscopy pulse rate and increased use of collimation and lower dose fluoroscopy modes. Conclusion: Hands-on fluoroscopy safety training substantially benefits from real-time patient and staff dosimetry incorporated into the class. Real-time dose display helps clinical staff visualize, internalize, and ultimately utilize the safety techniques learned during the training. RaySafe/Unfors/Fluke lent us a portable version of their RaySafe i2 Dosimetry System for 6 months.« less

  16. Reducing ionizing radiation doses during cardiac interventions in pregnant women.

    PubMed

    Orchard, Elizabeth; Dix, Sarah; Wilson, Neil; Mackillop, Lucy; Ormerod, Oliver

    2012-09-01

    There is concern over ionizing radiation exposure in women who are pregnant or of child-bearing age. Due to the increasing prevalence of congenital and acquired heart disease, the number of women who require cardiac interventions during pregnancy has increased. We have developed protocols for cardiac interventions in pregnant women and women of child-bearing age, aimed at substantially reducing both fluoroscopy duration and radiation doses. Over five years, we performed cardiac interventions on 15 pregnant women, nine postpartum women and four as part of prepregnancy assessment. Fluoroscopy times were minimized by simultaneous use of intracardiac echocardiography, and by using very low frame rates (2/second) during fluoroscopy. The procedures most commonly undertaken were closure of atrial septal defect (ASD) or patent foramen ovale (PFO) in 16 women, coronary angiograms in seven, right and left heart catheters in three and two stent placements. The mean screening time for all patients was 2.38 minutes (range 0.48-13.7), the median radiation dose was 66 (8.9-1501) Gy/cm(2). The median radiation dose to uterus was 1.92 (0.59-5.47) μGy, and the patient estimated dose was 0.24 (0.095-0.80) mSv. Ionizing radiation can be used safely in the management of severe cardiac structural disease in pregnancy, with very low ionizing radiation dose to the mother and extremely low exposure to the fetus. With experience, ionizing radiation doses at our institution have been reduced.

  17. CT fluoroscopy-guided percutaneous drainage: comparison of the one step and the Seldinger techniques.

    PubMed

    Kajiwara, Kenji; Yamagami, Takuji; Ishikawa, Masaki; Yoshimatsu, Rika; Baba, Yasutaka; Nakamura, Yuko; Fukumoto, Wataru; Awai, Kazuo

    2017-06-01

    To evaluate the one step technique compared with the Seldinger technique in computed tomography (CT) fluoroscopy-guided percutaneous drainage of abdominal and pelvic abscess. Seventy-six consecutive patients (49 men, 27 women; mean age 63.5 years, range 19-87 years) with abdominal and pelvic abscess were included in this study. Drainages were performed with the one step (n = 46) and with the Seldinger (n = 48) technique between September 2012 and June 2014. The technical success and clinical success rates were 95.8% and 93.5%, respectively, for the one step group, and 97.8% and 95.7%, respectively, for the Seldinger group. The mean procedure time was significantly shorter with the one step than with the Seldinger method (15.0 ± 4.3 min, range 10-29 min vs. 21.0 ± 9.5 min, range 13-54 min, p < .01). The mean abscess size and depth were 73.4 ± 44.0 mm and 42.5 ± 19.3 mm, respectively, in the one step group, and 61.0 ± 22.8 mm and 35.0 ± 20.7 mm in the Seldinger group. The one step technique was easier and faster than the Seldinger technique. The effectiveness of both techniques was similar for the CT fluoroscopy-guided percutaneous drainage of abdominal and pelvic abscess.

  18. Meta-Analysis of Zero or Near-Zero Fluoroscopy Use During Ablation of Cardiac Arrhythmias.

    PubMed

    Yang, Li; Sun, Ge; Chen, Xiaomei; Chen, Guangzhi; Yang, Shanshan; Guo, Ping; Wang, Yan; Wang, Dao Wen

    2016-11-15

    Data regarding the efficacy and safety of zero or near-zero fluoroscopic ablation of cardiac arrhythmias are limited. A literature search was conducted using PubMed and Embase for relevant studies through January 2016. Ten studies involving 2,261 patients were identified. Compared with conventional radiofrequency ablation method, zero or near-zero fluoroscopy ablation significantly showed reduced fluoroscopic time (standard mean difference [SMD] -1.62, 95% CI -2.20 to -1.05; p <0.00001), ablation time (SMD -0.16, 95% CI -0.29 to -0.04; p = 0.01), and radiation dose (SMD -1.94, 95% CI -3.37 to -0.51; p = 0.008). In contrast, procedure duration was not significantly different from that of conventional radiofrequency ablation (SMD -0.03, 95% CI -0.16 to 0.09; p = 0.58). There were no significant differences between both groups in immediate success rate (odds ratio [OR] 0.99, 95% CI 0.49 to 2.01; p = 0.99), long-term success rate (OR 1.13, 95% CI 0.42 to 3.02; p = 0.81), complication rates (OR 0.98, 95% CI 0.49 to 1.96; p = 0.95), and recurrence rates (OR 1.29, 95% CI 0.74 to 2.24; p = 0.37). In conclusion, radiation was significantly reduced in the zero or near-zero fluoroscopy ablation groups without compromising efficacy and safety. Copyright © 2016 Elsevier Inc. All rights reserved.

  19. Towards active image-guidance: tracking of a fiducial in the thorax during respiration under X-ray fluoroscopy

    NASA Astrophysics Data System (ADS)

    Siddique, Sami; Jaffray, David

    2007-03-01

    A central purpose of image-guidance is to assist the interventionalist with feedback of geometric performance in the direction of therapy delivery. Tradeoffs exist between accuracy, precision and the constraints imposed by parameters used in the generation of images. A framework that uses geometric performance as feedback to control these parameters can balance such tradeoffs in order to maintain the requisite localization precision for a given clinical procedure. We refer to this principle as Active Image-Guidance (AIG). This framework requires estimates of the uncertainty in the estimated location of the object of interest. In this study, a simple fiducial marker detected under X-ray fluoroscopy is considered and it is shown that a relation exists between the applied imaging dose and the uncertainty in localization for a given observer. A robust estimator of the location of a fiducial in the thorax during respiration under X-ray fluoroscopy is demonstrated using a particle filter based approach that outputs estimates of the location and the associated spatial uncertainty. This approach gives an rmse of 1.3mm and the uncertainty estimates are found to be correlated with the error in the estimates. Furthermore, the particle filtering approach is employed to output location estimates and the associated uncertainty not only at instances of pulsed exposure but also between exposures. Such a system has applications in image-guided interventions (surgery, radiotherapy, interventional radiology) where there are latencies between the moment of imaging and the act of intervention.

  20. Transforaminal Lumbar Interbody Fusion with Rigid Interspinous Process Fixation: A Learning Curve Analysis of a Surgeon Team's First 74 Cases.

    PubMed

    Doherty, Patrick; Welch, Arthur; Tharpe, Jason; Moore, Camille; Ferry, Chris

    2017-05-30

    Studies have shown that a significant learning curve may be associated with adopting minimally invasive transforaminal lumbar interbody fusion (MIS TLIF) with bilateral pedicle screw fixation (BPSF). Accordingly, several hybrid TLIF techniques have been proposed as surrogates to the accepted BPSF technique, asserting that less/fewer fixation(s) or less disruptive fixation may decrease the learning curve while still maintaining the minimally disruptive benefits. TLIF with interspinous process fixation (ISPF) is one such surrogate procedure. However, despite perceived ease of adaptability given the favorable proximity of the spinous processes, no evidence exists demonstrating whether or not the technique may possess its own inherent learning curve. The purpose of this study was to determine whether an intraoperative learning curve for one- and two-level TLIF + ISPF may exist for a single lead surgeon. Seventy-four consecutive patients who received one- or two-Level TLIF with rigid ISPF by a single lead surgeon were retrospectively reviewed. It was the first TLIF + ISPF case series for the lead surgeon. Intraoperative blood loss (EBL), hospitalization length-of-stay (LOS), fluoroscopy time, and postoperative complications were collected. EBL, LOS, and fluoroscopy time were modeled as a function of case number using multiple linear regression methods. A change point was included in each model to allow the trajectory of the outcomes to change during the duration of the case series. These change points were determined using profile likelihood methods. Models were fit using the maximum likelihood estimates for the change points. Age, sex, body mass index (BMI), and the number of treated levels were included as covariates. EBL, LOS, and fluoroscopy time did not significantly differ by age, sex, or BMI (p ≥ 0.12). Only EBL differed significantly by the number of levels (p = 0.026). The case number was not a significant predictor of EBL, LOS, or fluoroscopy time (p ≥ 0.21). At the time of data collection (mean time from surgery: 13.3 months), six patients had undergone revision due to interbody migration. No ISPF device complications were observed. Study outcomes support the ideal that TLIF + ISPF can be a readily adopted procedure without a significant intraoperative learning curve. However, the authors emphasize that further assessment of long-term healing outcomes is essential in fully characterizing both the efficacy and the indication learning curve for the TLIF + ISPF technique.

  1. Hubble Systems Optimize Hospital Schedules

    NASA Technical Reports Server (NTRS)

    2009-01-01

    Don Rosenthal, a former Ames Research Center computer scientist who helped design the Hubble Space Telescope's scheduling software, co-founded Allocade Inc. of Menlo Park, California, in 2004. Allocade's OnCue software helps hospitals reclaim unused capacity and optimize constantly changing schedules for imaging procedures. After starting to use the software, one medical center soon reported noticeable improvements in efficiency, including a 12 percent increase in procedure volume, 35 percent reduction in staff overtime, and significant reductions in backlog and technician phone time. Allocade now offers versions for outpatient and inpatient magnetic resonance imaging (MRI), ultrasound, interventional radiology, nuclear medicine, Positron Emission Tomography (PET), radiography, radiography-fluoroscopy, and mammography.

  2. Procedural techniques in sacral nerve modulation.

    PubMed

    Williams, Elizabeth R; Siegel, Steven W

    2010-12-01

    Sacral neuromodulation involves a staged process, including a screening trial and delayed formal implantation for those with substantial improvement. The advent of the tined lead has revolutionized the technology, allowing for a minimally invasive outpatient procedure to be performed under intravenous sedation. With the addition of fluoroscopy to the bilateral percutaneous nerve evaluation, there has been marked improvement in the placement of these temporary leads. Thus, the screening evaluation is now a better reflection of possible permanent improvement. Both methods of screening have advantages and disadvantages. Selection of a particular procedure should be tailored to individual patient characteristics. Subsequent implantation of the internal pulse generator (IPG) or explantation of an unsuccessful staged lead is straightforward outpatient procedure, providing minimal additional risk for the patient. Future refinement to the procedure may involve the introduction of a rechargeable battery, eliminating the need for IPG replacement at the end of the battery life.

  3. ICRP publication 121: radiological protection in paediatric diagnostic and interventional radiology.

    PubMed

    Khong, P-L; Ringertz, H; Donoghue, V; Frush, D; Rehani, M; Appelgate, K; Sanchez, R

    2013-04-01

    Paediatric patients have a higher average risk of developing cancer compared with adults receiving the same dose. The longer life expectancy in children allows more time for any harmful effects of radiation to manifest, and developing organs and tissues are more sensitive to the effects of radiation. This publication aims to provide guiding principles of radiological protection for referring clinicians and clinical staff performing diagnostic imaging and interventional procedures for paediatric patients. It begins with a brief description of the basic concepts of radiological protection, followed by the general aspects of radiological protection, including principles of justification and optimisation. Guidelines and suggestions for radiological protection in specific modalities - radiography and fluoroscopy, interventional radiology, and computed tomography - are subsequently covered in depth. The report concludes with a summary and recommendations. The importance of rigorous justification of radiological procedures is emphasised for every procedure involving ionising radiation, and the use of imaging modalities that are non-ionising should always be considered. The basic aim of optimisation of radiological protection is to adjust imaging parameters and institute protective measures such that the required image is obtained with the lowest possible dose of radiation, and that net benefit is maximised to maintain sufficient quality for diagnostic interpretation. Special consideration should be given to the availability of dose reduction measures when purchasing new imaging equipment for paediatric use. One of the unique aspects of paediatric imaging is with regards to the wide range in patient size (and weight), therefore requiring special attention to optimisation and modification of equipment, technique, and imaging parameters. Examples of good radiographic and fluoroscopic technique include attention to patient positioning, field size and adequate collimation, use of protective shielding, optimisation of exposure factors, use of pulsed fluoroscopy, limiting fluoroscopy time, etc. Major paediatric interventional procedures should be performed by experienced paediatric interventional operators, and a second, specific level of training in radiological protection is desirable (in some countries, this is mandatory). For computed tomography, dose reduction should be optimised by the adjustment of scan parameters (such as mA, kVp, and pitch) according to patient weight or age, region scanned, and study indication (e.g. images with greater noise should be accepted if they are of sufficient diagnostic quality). Other strategies include restricting multiphase examination protocols, avoiding overlapping of scan regions, and only scanning the area in question. Up-to-date dose reduction technology such as tube current modulation, organ-based dose modulation, auto kV technology, and iterative reconstruction should be utilised when appropriate. It is anticipated that this publication will assist institutions in encouraging the standardisation of procedures, and that it may help increase awareness and ultimately improve practices for the benefit of patients. Copyright © 2012. Published by Elsevier Ltd.

  4. The practical application of signal detection theory to image quality assessment in x-ray image intensifier-TV fluoroscopy.

    PubMed

    Marshall, N W

    2001-06-01

    This paper applies a published version of signal detection theory to x-ray image intensifier fluoroscopy data and compares the results with more conventional subjective image quality measures. An eight-bit digital framestore was used to acquire temporally contiguous frames of fluoroscopy data from which the modulation transfer function (MTF(u)) and noise power spectrum were established. These parameters were then combined to give detective quantum efficiency (DQE(u)) and used in conjunction with signal detection theory to calculate contrast-detail performance. DQE(u) was found to lie between 0.1 and 0.5 for a range of fluoroscopy systems. Two separate image quality experiments were then performed in order to assess the correspondence between the objective and subjective methods. First, image quality for a given fluoroscopy system was studied as a function of doserate using objective parameters and a standard subjective contrast-detail method. Following this, the two approaches were used to assess three different fluoroscopy units. Agreement between objective and subjective methods was good; doserate changes were modelled correctly while both methods ranked the three systems consistently.

  5. Persistent Atrial Fibrillation Ablation using the Tip-Versatile Ablation Catheter.

    PubMed

    Davies, Edward J; Clayton, Ben; Lines, Ian; Haywood, Guy A

    2016-07-01

    The mechanisms by which persistent atrial fibrillation (PsAF) develops are incompletely understood. Consequently, the optimal strategy for the ablative management of PsAF remains debated. Current methods are often time consuming, complex and non-reproducible. We assessed the Tip-Versatile Ablation Catheter (T-VAC) technique, a rapidly delivered, empirical technique based on the box-set concept using duty-cycled linear catheter ablation technology. Forty-four procedures in 40 patients undergoing PsAF ablation with the novel technique were prospectively entered onto a database: 27 de novo. Primary endpoint was freedom from arrhythmia at over two-year follow-up. Secondary endpoints were time to first arrhythmia recurrence, freedom from atrial fibrillation (AF) on and off antiarrhythmic drugs (AAD), procedural and fluoroscopy duration and complication rate. At mean follow-up of 33 months, absolute freedom from arrhythmia recurrence was 45% in the de novo group. Overall, at 33 (IQR 24-63) months, 60% of de novo patients were in sustained normal sinus rhythm and a further 15% reported only occasional paroxysms of AF at long-term follow-up. Procedure time was 192±25 mins, total energy delivered 2239±883s and fluoroscopy time was 60±10mins. In selected patients with persistent AF, a long-term rate of 60% arrhythmia free survival off AAD can be achieved using this novel T-VAC technique. Copyright © 2016 Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Published by Elsevier B.V. All rights reserved.

  6. Patient-specific rehearsal prior to EVAR: a pilot study.

    PubMed

    Desender, L; Rancic, Z; Aggarwal, R; Duchateau, J; Glenck, M; Lachat, M; Vermassen, F; Van Herzeele, I

    2013-06-01

    This study aims to evaluate feasibility, face validity, influence on technical factors and subjective sense of utility of patient-specific rehearsal (PsR) prior to endovascular aortic aneurysm repair (EVAR). A prospective, multicentre pilot study. Patients suitable for EVAR were enrolled and a three-dimensional (3D) model of the patient's anatomy was generated. Less than 24 h prior to the real case, rehearsals were conducted in the laboratory or clinical angiosuite. Technical metrics were recorded during both procedures. A subjective questionnaire was used to evaluate realism, technical and human factor aspects (scale 1-5). Ten patients were enrolled. In one case, the treatment plan was altered based on PsR. In 7/9 patients, the rehearsal significantly altered the optimal C-arm position for the proximal landing zone and an identical fluoroscopy angle was chosen in the real procedure. All team members found the rehearsal useful for selecting the optimal fluoroscopy angle (median 4). The realism of the EVAR procedure simulation was rated highly (median 4). All team members found the PsR useful to prepare the individual team members and the entire team (median 4). PsR for EVAR permits creation of realistic case studies. Subjective evaluation indicates that it may influence optimal C-arm angles and be valuable to prepare the entire team. A randomised controlled trial (RCT) is planned to evaluate how this technology may influence technical and team performance, ultimately leading to improved patient safety. Copyright © 2013 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

  7. Radiation Awareness for Endovascular Abdominal Aortic Aneurysm Repair in the Hybrid Operating Room. An Instant Patient Risk Chart for Daily Practice.

    PubMed

    de Ruiter, Quirina M; Gijsberts, Crystel M; Hazenberg, Constantijn E; Moll, Frans L; van Herwaarden, Joost A

    2017-06-01

    To determine which patient and C-arm characteristics are the strongest predictors of intraoperative patient radiation dose rates (DRs) during endovascular aneurysm repair (EVAR) procedures and create a patient risk chart. A retrospective analysis was performed of 74 EVAR procedures, including 16,889 X-ray runs using fixed C-arm imaging equipment. Four multivariate log-linear mixed models (with patient as a random effect) were constructed. Mean air kerma DR (DR AK , mGy/s) and the mean dose area product DR (DR DAP , mGycm 2 /s) were the outcome variables utilized for fluoroscopy as differentiated from digital subtraction angiography (DSA). These models were used to predict the maximum radiation duration allowed before a 2-Gy skin threshold (for DR AK ) or a 500-Gycm 2 threshold (for DR DAP ) was reached. The strongest predictor of DR AK and DR DAP for fluoroscopy imaging was the radiation protocol, with an increase of 200% when changing from "low" to "medium" and 410% from "low" to "normal." The strongest predictors of DR AK and DR DAP for DSA were C-arm angulation, with an increase of 47% per 30° of angulation, and body mass index (BMI), with an increase of 58% for every 5-point increase in BMI. Based on these models, a patient with a BMI of 30 kg/m 2 , combined with 45° of rotation and a field size of 800 cm 2 in the medium fluoroscopy protocol has a predicted DR AK of 0.39 mGy/s (or 85.5 minutes until the 2-Gy skin threshold is reached). While using comparable settings but switching the acquisition to a DSA with a "2 frames per second" protocol, the predicted DR AK will be 6.6 mGy/s (or 5.0 minutes until the 2-Gy threshold is reached). X-ray radiation DRs are constantly fluctuating during and between patients based on BMI, the protocols, C-arm position, and the image acquisitions that are used. An instant patient risk chart visualizes these radiation dose fluctuations and provides an overview of the expected duration of X-ray radiation, which can be used to predict when follow-up dose thresholds are reached during abdominal endovascular procedures.

  8. Real-time magnetic resonance-guided ablation of typical right atrial flutter using a combination of active catheter tracking and passive catheter visualization in man: initial results from a consecutive patient series.

    PubMed

    Hilbert, Sebastian; Sommer, Philipp; Gutberlet, Matthias; Gaspar, Thomas; Foldyna, Borek; Piorkowski, Christopher; Weiss, Steffen; Lloyd, Thomas; Schnackenburg, Bernhard; Krueger, Sascha; Fleiter, Christian; Paetsch, Ingo; Jahnke, Cosima; Hindricks, Gerhard; Grothoff, Matthias

    2016-04-01

    Recently cardiac magnetic resonance (CMR) imaging has been found feasible for the visualization of the underlying substrate for cardiac arrhythmias as well as for the visualization of cardiac catheters for diagnostic and ablation procedures. Real-time CMR-guided cavotricuspid isthmus ablation was performed in a series of six patients using a combination of active catheter tracking and catheter visualization using real-time MR imaging. Cardiac magnetic resonance utilizing a 1.5 T system was performed in patients under deep propofol sedation. A three-dimensional-whole-heart sequence with navigator technique and a fast automated segmentation algorithm was used for online segmentation of all cardiac chambers, which were thereafter displayed on a dedicated image guidance platform. In three out of six patients complete isthmus block could be achieved in the MR scanner, two of these patients did not need any additional fluoroscopy. In the first patient technical issues called for a completion of the procedure in a conventional laboratory, in another two patients the isthmus was partially blocked by magnetic resonance imaging (MRI)-guided ablation. The mean procedural time for the MR procedure was 109 ± 58 min. The intubation of the CS was performed within a mean time of 2.75 ± 2.21 min. Total fluoroscopy time for completion of the isthmus block ranged from 0 to 7.5 min. The combination of active catheter tracking and passive real-time visualization in CMR-guided electrophysiologic (EP) studies using advanced interventional hardware and software was safe and enabled efficient navigation, mapping, and ablation. These cases demonstrate significant progress in the development of MR-guided EP procedures. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2015. For permissions please email: journals.permissions@oup.com.

  9. TH-B-12A-01: TG124 “A Guide for Establishing a Credentialing and Privileging Program for Users of Fluoroscopic Equipment in Healthcare Organizations”

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Moore, M

    Fluoroscopy credentialing and privileging programs are being instituted because of recorded patient injuries and the widespread growth in fluoroscopy use by operators whose medical education did not include formal fluoroscopy training. This lack of training is recognized as a patient safety deficiency, and medical physicists and health physicists are finding themselves responsible for helping to establish fluoroscopy credentialing programs. While physicians are very knowledgeable about clinical credentials review and the privileging process, medical physicists and health physicists are not as familiar with the process and associated requirements. To assist the qualified medical physicist (QMP) and the radiation safety officer (RSO)more » with these new responsibilities, TG 124 provides an overview of the credentialing process, guidance for policy development and incorporating trained fluoroscopy users into a facility's established process, as well as recommendations for developing and maintaining a risk-based fluoroscopy safety training program. This lecture will review the major topics addressed in TG124 and relate them to practical situations. Learning Objectives: Understand the difference between credentialing and privileging. Understand the responsibilities, interaction and coordination among key individuals and committees. Understand options for integrating the QMP and/or RSO and Radiation Safety Committee into the credentialing and privileging process. Understand issues related to implementing the fluoroscopy safety training recommendations and with verifying and documenting successful completion.« less

  10. SU-C-18C-06: Radiation Dose Reduction in Body Interventional Radiology: Clinical Results Utilizing a New Imaging Acquisition and Processing Platform

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kohlbrenner, R; Kolli, KP; Taylor, A

    2014-06-01

    Purpose: To quantify the patient radiation dose reduction achieved during transarterial chemoembolization (TACE) procedures performed in a body interventional radiology suite equipped with the Philips Allura Clarity imaging acquisition and processing platform, compared to TACE procedures performed in the same suite equipped with the Philips Allura Xper platform. Methods: Total fluoroscopy time, cumulative dose area product, and cumulative air kerma were recorded for the first 25 TACE procedures performed to treat hepatocellular carcinoma (HCC) in a Philips body interventional radiology suite equipped with Philips Allura Clarity. The same data were collected for the prior 85 TACE procedures performed to treatmore » HCC in the same suite equipped with Philips Allura Xper. Mean values from these cohorts were compared using two-tailed t tests. Results: Following installation of the Philips Allura Clarity platform, a 42.8% reduction in mean cumulative dose area product (3033.2 versus 1733.6 mGycm∧2, p < 0.0001) and a 31.2% reduction in mean cumulative air kerma (1445.4 versus 994.2 mGy, p < 0.001) was achieved compared to similar procedures performed in the same suite equipped with the Philips Allura Xper platform. Mean total fluoroscopy time was not significantly different between the two cohorts (1679.3 versus 1791.3 seconds, p = 0.41). Conclusion: This study demonstrates a significant patient radiation dose reduction during TACE procedures performed to treat HCC after a body interventional radiology suite was converted to the Philips Allura Clarity platform from the Philips Allura Xper platform. Future work will focus on evaluation of patient dose reduction in a larger cohort of patients across a broader range of procedures and in specific populations, including obese patients and pediatric patients, and comparison of image quality between the two platforms. Funding for this study was provided by Philips Healthcare, with 5% salary support provided to authors K. Pallav Kolli and Robert G. Gould for time devoted to the study. Data acquisition and analysis was performed by the authors independent of the funding source.« less

  11. Retrieval of detached fragment of central venous pressure catheter (CVP) lodged in the right ventricle and pulmonary artery: a case report.

    PubMed

    Sakijan, A S; Zambahari, R; Annuar, Z; Yahya, O; Ali, J

    1990-12-01

    A successful retrieval of a detached segment of a CVP catheter by percutaneous right transfemoral venous route, using a Dotter intravascular retriever basket, is reported. The procedure was monitored under fluoroscopy. Only local anaesthesia, which was infiltrated around the puncture site, was given to the patient. No significant complication was encountered. Successful retrieval of the detached catheter fragment by percutaneous means obviates the need for thoracotomy.

  12. Novel Image-Guided Management of a Uterine Arteriovenous Malformation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Przybojewski, Stefan J., E-mail: drstefanp@hotmail.com; Sadler, David J.

    The investigators present a novel image-guided embolization, not previously described, of a uterine arteriovenous malformation (AVM) resistant to endovascular management. The uterus was exposed surgically, and Histoacryl (Braun, Fulda, Germany) was injected directly into the nidus using ultrasound guidance and fluoroscopy. The patient had a successful full-term pregnancy after this procedure. This technique may be a useful alternative management strategy in patients with uterine AVM who fail traditional endovascular embolization and who still desire fertility.

  13. Collection of DICOM RDSR (Digital Imaging and Communication in Medicine, Radiation Dose Structured Report) Information Aimed at Reducing Patient Exposure Dose.

    PubMed

    Morota, Koichi; Moritake, Takashi; Sun, Lue; Ishihara, Takahiro; Kuma, Natsuyo; Murata, Satomi; Yamada, Takahiro; Okazaki, Ryuji

    2016-01-01

    The recent progress in angiography technology bestows benefits on patients for minimally invasive than surgery, while there has been an increase in the number of cases involving stochastic effects, such as radiation dermatitis, resulting from upgrading of the procedure because of an extension of the time for fluoroscopy and the number of shots. Recent CT equipment saves the dose data along with image data about the information management for patient exposure dose, which is used for management of individual cumulative dose and the presumed effective dose, using digital imaging and communication in medicine (DICOM). We extracted detailed information about shooting conditions and dose from the DICOM radiation dose structured report (DICOM RDSR) in the angiography area, and evaluated the trend of patient exposure dose in each procedure. As a result, we found that cases exceeding 3 Gy which needed observation in the head region were 16.7% and in the heart region were 27.3%. We also found that angiography had a higher dose of shooting than did fluoroscopy, and that the diagnosis and treatment with tumor involvement required a exposure dose than did vascular lesion. In this paper, we review the shooting conditions as a root of DICOM RDSR information and consider the possibility of planning for further reduction of the exposure dose.

  14. Navigational Guidance and Ablation Planning Tools for Interventional Radiology.

    PubMed

    Sánchez, Yadiel; Anvari, Arash; Samir, Anthony E; Arellano, Ronald S; Prabhakar, Anand M; Uppot, Raul N

    Image-guided biopsy and ablation relies on successful identification and targeting of lesions. Currently, image-guided procedures are routinely performed under ultrasound, fluoroscopy, magnetic resonance imaging, or computed tomography (CT) guidance. However, these modalities have their limitations including inadequate visibility of the lesion, lesion or organ or patient motion, compatibility of instruments in an magnetic resonance imaging field, and, for CT and fluoroscopy cases, radiation exposure. Recent advances in technology have resulted in the development of a new generation of navigational guidance tools that can aid in targeting lesions for biopsy or ablations. These navigational guidance tools have evolved from simple hand-held trajectory guidance tools, to electronic needle visualization, to image fusion, to the development of a body global positioning system, to growth in cone-beam CT, and to ablation volume planning. These navigational systems are promising technologies that not only have the potential to improve lesion targeting (thereby increasing diagnostic yield of a biopsy or increasing success of tumor ablation) but also have the potential to decrease radiation exposure to the patient and staff, decrease procedure time, decrease the sedation requirements, and improve patient safety. The purpose of this article is to describe the challenges in current standard image-guided techniques, provide a definition and overview for these next-generation navigational devices, and describe the current limitations of these, still evolving, next-generation navigational guidance tools. Copyright © 2017 Elsevier Inc. All rights reserved.

  15. Non-photorealistic rendering of virtual implant models for computer-assisted fluoroscopy-based surgical procedures

    NASA Astrophysics Data System (ADS)

    Zheng, Guoyan

    2007-03-01

    Surgical navigation systems visualize the positions and orientations of surgical instruments and implants as graphical overlays onto a medical image of the operated anatomy on a computer monitor. The orthopaedic surgical navigation systems could be categorized according to the image modalities that are used for the visualization of surgical action. In the so-called CT-based systems or 'surgeon-defined anatomy' based systems, where a 3D volume or surface representation of the operated anatomy could be constructed from the preoperatively acquired tomographic data or through intraoperatively digitized anatomy landmarks, a photorealistic rendering of the surgical action has been identified to greatly improve usability of these navigation systems. However, this may not hold true when the virtual representation of surgical instruments and implants is superimposed onto 2D projection images in a fluoroscopy-based navigation system due to the so-called image occlusion problem. Image occlusion occurs when the field of view of the fluoroscopic image is occupied by the virtual representation of surgical implants or instruments. In these situations, the surgeon may miss part of the image details, even if transparency and/or wire-frame rendering is used. In this paper, we propose to use non-photorealistic rendering to overcome this difficulty. Laboratory testing results on foamed plastic bones during various computer-assisted fluoroscopybased surgical procedures including total hip arthroplasty and long bone fracture reduction and osteosynthesis are shown.

  16. Urethral Stenting for Obstructive Uropathy Utilizing Digital Radiography for Guidance: Feasibility and Clinical Outcome in 26 Dogs.

    PubMed

    Radhakrishnan, A

    2017-03-01

    Urethral stent placement is an interventional treatment option to alleviate urethral outflow obstruction. It has been described utilizing fluoroscopy, but fluoroscopy is not as readily available in private practice as digital radiography. To describe the use of digital radiography for urethral stent placement in dogs with obstructive uropathy. Twenty-six client-owned dogs presented for dysuria associated with benign and malignant causes of obstructive uropathy that underwent urethral stent placement. Retrospective study. Causes of obstructive uropathy included transitional cell carcinoma, prostatic carcinoma, hemangiosarcoma, obstructive proliferative urethritis, compressive vaginal leiomyosarcoma, and detrusor-sphincter dyssynergia. Survival time range was 1-48 months (median, 5 months). All dogs were discharged from the hospital with urine outflow restored. Intraprocedural complications included guide wire penetration of the urethral wall in 1 dog and improper stent placement in a second dog. Both complications were successfully managed at the time of the procedure with no follow-up problems noted in either patient. Urethral stent placement can be successfully performed utilizing digital radiography. The complications experienced can be avoided by more cautious progression with each step through the procedure and serial radiography. The application of digital radiography may allow treatment of urethral obstruction to become more readily available. Copyright © 2017 The Authors. Journal of Veterinary Internal Medicine published by Wiley Periodicals, Inc. on behalf of the American College of Veterinary Internal Medicine.

  17. Effective Dose of CT- and Fluoroscopy-Guided Perineural/Epidural Injections of the Lumbar Spine: A Comparative Study

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Schmid, Gebhard; Schmitz, Alexander; Borchardt, Dieter

    The objective of this study was to compare the effective radiation dose of perineural and epidural injections of the lumbar spine under computed tomography (CT) or fluoroscopic guidance with respect to dose-reduced protocols. We assessed the radiation dose with an Alderson Rando phantom at the lumbar segment L4/5 using 29 thermoluminescence dosimeters. Based on our clinical experience, 4-10 CT scans and 1-min fluoroscopy are appropriate. Effective doses were calculated for CT for a routine lumbar spine protocol and for maximum dose reduction; as well as for fluoroscopy in a continuous and a pulsed mode (3-15 pulses/s). Effective doses under CTmore » guidance were 1.51 mSv for 4 scans and 3.53 mSv for 10 scans using a standard protocol and 0.22 mSv and 0.43 mSv for the low-dose protocol. In continuous mode, the effective doses ranged from 0.43 to 1.25 mSv for 1-3 min of fluoroscopy. Using 1 min of pulsed fluoroscopy, the effective dose was less than 0.1 mSv for 3 pulses/s. A consequent low-dose CT protocol reduces the effective dose compared to a standard lumbar spine protocol by more than 85%. The latter dose might be expected when applying about 1 min of continuous fluoroscopy for guidance. A pulsed mode further reduces the effective dose of fluoroscopy by 80-90%.« less

  18. Radiation dose and image quality for paediatric interventional cardiology

    NASA Astrophysics Data System (ADS)

    Vano, E.; Ubeda, C.; Leyton, F.; Miranda, P.

    2008-08-01

    Radiation dose and image quality for paediatric protocols in a biplane x-ray system used for interventional cardiology have been evaluated. Entrance surface air kerma (ESAK) and image quality using a test object and polymethyl methacrylate (PMMA) phantoms have been measured for the typical paediatric patient thicknesses (4-20 cm of PMMA). Images from fluoroscopy (low, medium and high) and cine modes have been archived in digital imaging and communications in medicine (DICOM) format. Signal-to-noise ratio (SNR), figure of merit (FOM), contrast (CO), contrast-to-noise ratio (CNR) and high contrast spatial resolution (HCSR) have been computed from the images. Data on dose transferred to the DICOM header have been used to test the values of the dosimetric display at the interventional reference point. ESAK for fluoroscopy modes ranges from 0.15 to 36.60 µGy/frame when moving from 4 to 20 cm PMMA. For cine, these values range from 2.80 to 161.10 µGy/frame. SNR, FOM, CO, CNR and HCSR are improved for high fluoroscopy and cine modes and maintained roughly constant for the different thicknesses. Cumulative dose at the interventional reference point resulted 25-45% higher than the skin dose for the vertical C-arm (depending of the phantom thickness). ESAK and numerical image quality parameters allow the verification of the proper setting of the x-ray system. Knowing the increases in dose per frame when increasing phantom thicknesses together with the image quality parameters will help cardiologists in the good management of patient dose and allow them to select the best imaging acquisition mode during clinical procedures.

  19. Novel System for Real-Time Integration of 3-D Echocardiography and Fluoroscopy for Image-Guided Cardiac Interventions: Preclinical Validation and Clinical Feasibility Evaluation.

    PubMed

    Arujuna, Aruna V; Housden, R James; Ma, Yingliang; Rajani, Ronak; Gao, Gang; Nijhof, Niels; Cathier, Pascal; Bullens, Roland; Gijsbers, Geert; Parish, Victoria; Kapetanakis, Stamatis; Hancock, Jane; Rinaldi, C Aldo; Cooklin, Michael; Gill, Jaswinder; Thomas, Martyn; O'neill, Mark D; Razavi, Reza; Rhode, Kawal S

    2014-01-01

    Real-time imaging is required to guide minimally invasive catheter-based cardiac interventions. While transesophageal echocardiography allows for high-quality visualization of cardiac anatomy, X-ray fluoroscopy provides excellent visualization of devices. We have developed a novel image fusion system that allows real-time integration of 3-D echocardiography and the X-ray fluoroscopy. The system was validated in the following two stages: 1) preclinical to determine function and validate accuracy; and 2) in the clinical setting to assess clinical workflow feasibility and determine overall system accuracy. In the preclinical phase, the system was assessed using both phantom and porcine experimental studies. Median 2-D projection errors of 4.5 and 3.3 mm were found for the phantom and porcine studies, respectively. The clinical phase focused on extending the use of the system to interventions in patients undergoing either atrial fibrillation catheter ablation (CA) or transcatheter aortic valve implantation (TAVI). Eleven patients were studied with nine in the CA group and two in the TAVI group. Successful real-time view synchronization was achieved in all cases with a calculated median distance error of 2.2 mm in the CA group and 3.4 mm in the TAVI group. A standard clinical workflow was established using the image fusion system. These pilot data confirm the technical feasibility of accurate real-time echo-fluoroscopic image overlay in clinical practice, which may be a useful adjunct for real-time guidance during interventional cardiac procedures.

  20. Thyroid Dose During Neurointerventional Procedures: Does Lead Shielding Reduce the Dose?

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Shortt, C. P.; Fanning, N. F.; Malone, L.

    2007-09-15

    Purpose. To assess radiation dose to the thyroid in patients undergoing neurointerventional procedures and to evaluate dose reduction to the thyroid by lead shielding. Methods and Materials. A randomized patient study was undertaken to evaluate the dose reduction by thyroid lead shields and assess their practicality in a clinical setting. Sixty-five patients attending for endovascular treatment of arteriovenous malformations (AVMs) and aneurysms were randomized into one of 2 groups a) No Thyroid Shield and b) Thyroid Lead Shield. Two thermoluminescent dosimeters (TLDs) were placed over the thyroid gland (1 on each side) at constant positions on each patient in bothmore » groups. A thyroid lead shield (Pb eq. 0.5 mm) was placed around the neck of patients in the thyroid lead shield group after the neurointerventional radiologist had obtained satisfactory working access above the neck. The total dose-area-product (DAP) value, number and type of digital subtraction angiography (DSA) runs and fluoroscopy time were recorded for all patients. Results. Of the 72 patients who initially attended for neurointerventional procedures, 7 were excluded due to failure to consent or because of procedures involving access to the external carotid circulation. Of the remaining 65 who were randomized, a further 9 were excluded due to; procedureabandonment, unfeasible shield placement or shield interference with the procedure. Patient demographics included mean age of 47.9 yrs (15-74), F:M=1.4:1. Mean fluoroscopy time was 25.9 min. Mean DAP value was 13,134.8 cGy.cm{sup 2} and mean number of DSA runs was 13.4. The mean relative thyroid doses were significantly different (p< 0.001) between the unshielded (7.23 mSv/cGy2 x 105) and shielded groups (3.77 mSv/cGy2 x 105). A mean thyroid dose reduction of 48% was seen in the shielded group versus the unshielded group. Conclusion. Considerable doses to the thyroid are incurred during neurointerventional procedures, highlighting the need for increased awareness of patient radiation protection. Thyroid lead shielding yields significant radiation protection, is inexpensive and when not obscuring the field of view, should be used routinely.« less

  1. Three-dimensional rotational angiography fused with multimodal imaging modalities for targeted endomyocardial injections in the ischaemic heart.

    PubMed

    Dauwe, Dieter Frans; Nuyens, Dieter; De Buck, Stijn; Claus, Piet; Gheysens, Olivier; Koole, Michel; Coudyzer, Walter; Vanden Driessche, Nina; Janssens, Laurens; Ector, Joris; Dymarkowski, Steven; Bogaert, Jan; Heidbuchel, Hein; Janssens, Stefan

    2014-08-01

    Biological therapies for ischaemic heart disease require efficient, safe, and affordable intramyocardial delivery. Integration of multiple imaging modalities within the fluoroscopy framework can provide valuable information to guide these procedures. We compared an anatomo-electric method (LARCA) with a non-fluoroscopic electromechanical mapping system (NOGA(®)). LARCA integrates selective three-dimensional-rotational angiograms with biplane fluoroscopy. To identify the infarct region, we studied LARCA-fusion with pre-procedural magnetic resonance imaging (MRI), dedicated CT, or (18)F-FDG-PET/CT. We induced myocardial infarction in 20 pigs by 90-min LAD occlusion. Six weeks later, we compared peri-infarct delivery accuracy of coloured fluospheres using sequential NOGA(®)- and LARCA-MRI-guided vs. LARCA-CT- and LARCA-(18)F-FDG-PET/CT-guided intramyocardial injections. MRI after 6 weeks revealed significant left ventricular (LV) functional impairment and remodelling (LVEF 31 ± 3%, LVEDV 178 ± 15 mL, infarct size 17 ± 2% LV mass). During NOGA(®)-procedures, three of five animals required DC-shock for major ventricular arrhythmias vs. one of ten during LARCA-procedures. Online procedure time was shorter for LARCA than NOGA(®) (77 ± 6 vs. 130 ± 3 min, P < 0.0001). Absolute distance of injection spots to the infarct border was similar for LARCA-MRI (4.8 ± 0.5 mm) and NOGA(®) (5.4 ± 0.5 mm). LARCA-CT-integration allowed closer approximation of the targeted border zone than LARCA-PET (4.0 ± 0.5 mm vs. 6.2 ± 0.6 mm, P < 0.05). Three-dimensional -rotational angiography fused with multimodal imaging offers a new, cost-effective, and safe strategy to guide intramyocardial injections. Endoventricular procedure times and arrhythmias compare favourably to NOGA(®), without compromising injection accuracy. LARCA-based fusion imaging is a promising enabling technology for cardiac biological therapies. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2014. For permissions please email: journals.permissions@oup.com.

  2. Performance evaluation of image-intensifier-TV fluoroscopy systems

    NASA Astrophysics Data System (ADS)

    van der Putten, Wilhelm J.; Bouley, Shawn

    1995-05-01

    Through use of a computer model and an aluminum low contrast phantom developed in-house, a method has been developed which is able to grade the imaging performance of fluoroscopy systems through use of a variable, K. This parameter was derived from Rose's model of image perception and is here used as a figure of merit to grade fluoroscopy systems. From Rose's model for an ideal system, a typical value of K for the perception of low contrast details should be between 3 and 7, assuming threshold vision by human observers. Thus, various fluoroscopy systems are graded with different values of K, with a lower value of K indicating better imaging performance of the system. A series of fluoroscopy systems have been graded where the best system produces a value in the low teens, while the poorest systems produce a value in the low twenties. Correlation with conventional image quality measurements is good and the method has the potential for automated assessment of image quality.

  3. Magnetic resonance venography and three-dimensional image fusion guidance provide a novel paradigm for endovascular recanalization of chronic central venous occlusion.

    PubMed

    Schwein, Adeline; Lu, Tony; Chinnadurai, Ponraj; Kitkungvan, Danai; Shah, Dipan J; Chakfe, Nabil; Lumsden, Alan B; Bismuth, Jean

    2017-01-01

    Endovascular recanalization is considered first-line therapy for chronic central venous occlusion (CVO). Unlike arteries, in which landmarks such as wall calcifications provide indirect guidance for endovascular navigation, sclerotic veins without known vascular branching patterns impose significant challenges. Therefore, safe wire access through such chronic lesions mostly relies on intuition and experience. Studies have shown that magnetic resonance venography (MRV) can be performed safely in these patients, and the boundaries of occluded veins may be visualized on specific MRV sequences. Intraoperative image fusion techniques have become more common to guide complex arterial endovascular procedures. The aim of this study was to assess the feasibility and utility of MRV and intraoperative cone-beam computed tomography (CBCT) image fusion technique during endovascular CVO recanalization. During the study period, patients with symptomatic CVO and failed standard endovascular recanalization underwent further recanalization attempts with use of intraoperative MRV image fusion guidance. After preoperative MRV and intraoperative CBCT image coregistration, a virtual centerline path of the occluded segment was electronically marked in MRV and overlaid on real-time two-dimensional fluoroscopy images. Technical success, fluoroscopy times, radiation doses, number of venograms before recanalization, and accuracy of the virtual centerline overlay were evaluated. Four patients underwent endovascular CVO recanalization with use of intraoperative MRV image fusion guidance. Mean (± standard deviation) time for image fusion was 6:36 ± 00:51 mm:ss. The lesion was successfully crossed in all patients without complications. Mean fluoroscopy time for lesion crossing was 12.5 ± 3.4 minutes. Mean total fluoroscopy time was 28.8 ± 6.5 minutes. Mean total radiation dose was 15,185 ± 7747 μGy/m 2 , and mean radiation dose from CBCT acquisition was 2788 ± 458 μGy/m 2 (18% of mean total radiation dose). Mean number of venograms before recanalization was 1.6 ± 0.9, whereas two lesions were crossed without any prior venography. On qualitative analysis, virtual centerlines from MRV were aligned with actual guidewire trajectory on fluoroscopy in all four cases. MRV image fusion is feasible and may improve success, safety, and the surgeon's confidence during CVO recanalization. Similar to arterial interventions, three-dimensional MRV imaging and image fusion techniques could foster innovative solutions for such complex venous interventions and have the potential to affect a great number of patients. Copyright © 2016 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

  4. Percutaneous Steerable Robotic Tool Delivery Platform and Metal MEMS Device for Tissue Manipulation and Approximation: Closure of Patent Foramen Ovale in an Animal Model

    PubMed Central

    Vasilyev, Nikolay V.; Gosline, Andrew H.; Butler, Evan; Lang, Nora; Codd, Patrick J.; Yamauchi, Haruo; Feins, Eric N.; Folk, Chris R.; Cohen, Adam L.; Chen, Richard; Zurakowski, David; del Nido, Pedro J.; Dupont, Pierre E

    2013-01-01

    Background Beating-heart image-guided intracardiac interventions have been evolving rapidly. To extend the domain of catheter-based and transcardiac interventions into reconstructive surgery, a new robotic tool delivery platform (TDP) and tissue approximation device have been developed. Initial results employing these tools to perform patent foramen ovale (PFO) closure are described. Methods and Results A robotic TDP comprised of superelastic metal tubes provides the capability of delivering and manipulating tools and devices inside the beating heart. A new device technology is also presented that utilizes a metal-based MicroElectroMechanical Systems (MEMS) manufacturing process to produce fully-assembled and fully-functional millimeter-scale tools. As a demonstration of both technologies, a PFO creation and closure was performed in a swine model. In the first group of animals (N=10), a preliminary study was performed. The procedural technique was validated with a transcardiac handheld delivery platform and epicardial echocardiography, video-assisted cardioscopy and fluoroscopy. In the second group (N=9), the procedure was performed percutaneously using the robotic TDP under epicardial echocardiography and fluoroscopy imaging. All PFO’s were completely closed in the first group. In the second group, the PFO was not successfully created in 1 animal, and the defects were completely closed in 6 of the 8 remaining animals. Conclusions In contrast to existing robotic catheter technologies, the robotic TDP utilizes a combination of stiffness and active steerability along its length to provide the positioning accuracy and force application capability necessary for tissue manipulation. In combination with a MEMS tool technology, it can enable reconstructive procedures inside the beating heart. PMID:23899870

  5. Robotic-Assisted Versus Manual Prostatic Arterial Embolization for Benign Prostatic Hyperplasia: A Comparative Analysis.

    PubMed

    Bagla, Sandeep; Smirniotopoulos, John; Orlando, Julie C; Piechowiak, Rachel

    2017-03-01

    Prostatic artery embolization (PAE) is a safe and efficacious procedure for benign prostatic hyperplasia (BPH), though is technically challenging. We present our experience of technical and clinical outcomes of robotic and manual PAE in patients with BPH. IRB-approved retrospective study of 40 consecutive patients 49-81 years old with moderate or severe grade BPH from May 2014 to July 2015: 20 robotic-assisted PAE (group 1), 20 manual PAE (group 2). Robotic-assisted PAE was performed using the Magellan Robotic System. American Urological Association (AUA-SI) score, cost, technical and clinical success, radiation dose, fluoroscopy, and procedure time were reviewed. Statistical analysis was performed within and between each group using paired t test and one-way analysis of variance respectively, at 1 and 3 months. No significant baseline differences in age and AUA-SI between groups. Technical success was 100% (group 1) and 95% (group 2). One unsuccessful subject from group 2 returned for a successful embolization using robotic assistance. Fluoroscopy and procedural times were similar between groups, with a non-significant lower patient radiation dose in group 1 (30,632.8 mGy/cm 2 vs 35,890.9, p = 0.269). Disposable cost was significantly different between groups with the robotic-assisted PAE incurring a higher cost (group 1 $4530.2; group 2 $1588.5, p < 0.0001). Clinical improvement was significant in both arms at 3 months: group 1 mean change in AUA-SI of 8.3 (p = 0.006), group 2: 9.6 (p < 0.0001). No minor or major complications occurred. Robotic-assisted PAE offers technical success comparable to manual PAE, with similar clinical improvement with an increased cost.

  6. Efficacy and safety of the second-generation cryoballoons versus radiofrequency ablation for the treatment of paroxysmal atrial fibrillation: a systematic review and meta-analysis.

    PubMed

    Jiang, Jingbo; Li, Jinyi; Zhong, Guoqiang; Jiang, Junjun

    2017-01-01

    Currently, radiofrequency (RF) and cryoballoon are the most commonly used ablation technologies for atrial fibrillation (AF). We performed a meta-analysis to assess the efficacy and safety of the second-generation cryoballoons (CB-2) compared with RF for paroxysmal atrial fibrillation (PAF) ablation. The PubMed, Cochrane Library, and Embase databases were searched and qualified studies were identified. The primary clinical outcome was the recurrence rate of atrial tachyarrhythmia (AT), and the secondary clinical outcomes were procedure time, fluoroscopy time, and the complications that followed. Nine observational studies (2336 patients) with a mean follow-up period ranging from 8.8 to 16.8 months were included. The CB-2 group was associated with a significantly lower recurrence rate of ATs (20.8 versus 29.8 %, p = 0.01). In subgroup analysis, compared with non-contact force sensing (NCF) catheter, using CB-2 showed significantly reduced incidence of ATs (22.0 versus 38.5 %, p < 0.00001). However, the difference became negligible in contrast with contact force sensing (CF) catheter. Moreover, the CB-2 group had a tendency to decrease procedure time (weighted mean difference -39.72 min, p = 0.0003), whereas fluoroscopy time was similar between the two groups. The total complication rate showed no statistical difference (8.8 versus 4.4 %, p = 0.08). Almost all the cases of phrenic nerve palsy occurred in the CB-2 group, whereas pericardial tamponade was seldom manifested in the CB-2 group. CB-2 tended to be more effective in comparison to NCF catheter and at least non-inferior to CF catheter, with shorter procedure time and similar safety endpoint.

  7. Comparison of local versus general anesthesia in patients undergoing transcatheter aortic valve replacement: A meta-analysis.

    PubMed

    Villablanca, Pedro A; Mohananey, Divyanshu; Nikolic, Katarina; Bangalore, Sripal; Slovut, David P; Mathew, Verghese; Thourani, Vinod H; Rode's-Cabau, Josep; Núñez-Gil, Iván J; Shah, Tina; Gupta, Tanush; Briceno, David F; Garcia, Mario J; Gutsche, Jacob T; Augoustides, John G; Ramakrishna, Harish

    2018-02-01

    Transcatheter aortic valve replacement (TAVR) is typically performed under general anesthesia (GA). However, there is increasing data supporting the safety of performing TAVR under local anesthesia/conscious sedation (LA). We performed a meta-analysis to gain better understanding of the safety and efficacy of LA versus GA in patients with severe aortic stenosis undergoing TAVR. We comprehensively searched EMBASE, PubMed, and Web of Science. Effect sizes were summarized using risk ratios (RRs) difference of the mean (DM), and 95% CIs (confidence intervals) for dichotomous and continuous variables respectively. Twenty-six studies and 10,572 patients were included in the meta-analysis. The use of LA for TAVR was associated with lower overall 30-day mortality (RR, 0.73; 95% CI, 0.57-0.93; P = 0.01), use of inotropic/vasopressor drugs (RR, 0.45; 95% CI, 0.28-0.72; P < 0.001), hospital length of stay (LOS) (DM, -2.09; 95% CI, -3.02 to -1.16; P < 0.001), intensive care unit LOS (DM, -0.18; 95% CI, -0.31 to -0.04; P = 0.01), procedure time (DM, -25.02; 95% CI, -32.70 to -17.35; P < 0.001); and fluoroscopy time (DM, -1.63; 95% CI, -3.02 to -0.24; P = 0.02). No differences were observed between LA and GA for stroke, cardiovascular mortality, myocardial infarction, permanent pacemaker implantation, acute kidney injury, paravalvular leak, vascular complications, major bleeding, procedural success, conduction abnormalities, and annular rupture. Our meta-analysis suggests that use of LA for TAVR is associated with a lower 30-day mortality, shorter procedure time, fluoroscopy time, ICU LOS, hospital length of stay, and reduced need for inotropic support. © 2017 Wiley Periodicals, Inc.

  8. Sheathless guide catheter in transradial percutaneous coronary intervention for ST-segment elevation myocardial infarction.

    PubMed

    Miyasaka, Masaki; Tada, Norio; Kato, Shigeaki; Kami, Masahiro; Horie, Kazunori; Honda, Taku; Takizawa, Kaname; Otomo, Tatsushi; Inoue, Naoto

    2016-05-01

    The aim of this study was to assess the safety and efficacy of sheathless guide catheters in transradial percutaneous coronary intervention (PCI) for ST-segment elevation myocardial infarction (STEMI). Transradial PCI for STEMI offers significant clinical benefits, including a reduced incidence of vascular complications. As the size of the radial artery is small, the radial artery is frequently damaged in this procedure using large-bore catheters. A sheathless guide catheter offers a solution to this problem as it does not require an introducer sheath. However, the efficacy and safety of sheathless guide catheters remain to be fully determined in emergent transradial PCI for STEMI. Data on consecutive STEMI patients undergoing primary PCI at the Sendai Kousei Hospital between September 2010 and May 2013 were analyzed. The primary endpoint was the rate of acute procedural success without access site crossover. Secondary endpoints included door-to-balloon time, fluoroscopy time, volume of contrast, and radial artery stenosis or occlusion rate. We conducted transradial PCI for 478 patients with STEMI using a sheathless guide catheter. Acute procedural success was achieved in 466 patients (97.5%). The median door-to-balloon time was 45 min (range, 15-317 min). The median fluoroscopy time was 16.4 min (range, 10-90 min). The median volume of contrast was 134 mL (range, 31-431 mL). Radial stenosis or occlusion developed in 14 (3.8%) of the 370 evaluable patients. This study showed that use of a sheathless guide catheter taking a transradial approach was effective and safe in primary PCI for STEMI. © 2015 Wiley Periodicals, Inc. © 2015 Wiley Periodicals, Inc.

  9. Quality-of-Life Assessment After Palliative Interventions to Manage Malignant Ureteral Obstruction

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Monsky, Wayne Laurence, E-mail: wemonsky@msn.com; Molloy, Chris; Jin, Bedro

    2013-10-15

    Purpose: Malignancies may cause urinary tract obstruction, which is often relieved with placement of a percutaneous nephrostomy tube, an internal double J nephro-ureteric stent (double J), or an internal external nephroureteral stent (NUS). We evaluated the affect of these palliative interventions on quality of life (QoL) using previously validated surveys. Methods: Forty-six patients with malignancy related ureteral obstruction received nephrostomy tubes (n = 16), double J stents (n = 15), or NUS (n = 15) as determined by a multidisciplinary team. QoL surveys were administered at 7, 30, and 90 days after the palliative procedure to evaluate symptoms and physical,more » social, functional, and emotional well-being. Number of related procedures, fluoroscopy time, and complications were documented. Kruskal-Wallis and Friedman's test were used to compare patients at 7, 30, and 90 days. Spearman's rank correlation coefficient was used to assess correlations between clinical outcomes/symptoms and QoL. Results: Responses to QoL surveys were not significantly different for patients receiving nephrostomies, double J stents, or NUS at 7, 30, or 90 days. At 30 and 90 days there were significantly higher reported urinary symptoms and pain in those receiving double J stents compared with nephrostomies (P = 0.0035 and P = 0.0189, respectively). Significantly greater fluoroscopy time was needed for double J stent-related procedures (P = 0.0054). Nephrostomy tubes were associated with more frequent minor complications requiring additional changes. Conclusion: QoL was not significantly different. However, a greater incidence of pain in those receiving double J stents and more frequent tube changes in those with nephrostomy tubes should be considered when choosing palliative approaches.« less

  10. Advanced electrophysiologic mapping systems: an evidence-based analysis.

    PubMed

    2006-01-01

    To assess the effectiveness, cost-effectiveness, and demand in Ontario for catheter ablation of complex arrhythmias guided by advanced nonfluoroscopy mapping systems. Particular attention was paid to ablation for atrial fibrillation (AF). Tachycardia Tachycardia refers to a diverse group of arrhythmias characterized by heart rates that are greater than 100 beats per minute. It results from abnormal firing of electrical impulses from heart tissues or abnormal electrical pathways in the heart because of scars. Tachycardia may be asymptomatic, or it may adversely affect quality of life owing to symptoms such as palpitations, headaches, shortness of breath, weakness, dizziness, and syncope. Atrial fibrillation, the most common sustained arrhythmia, affects about 99,000 people in Ontario. It is associated with higher morbidity and mortality because of increased risk of stroke, embolism, and congestive heart failure. In atrial fibrillation, most of the abnormal arrhythmogenic foci are located inside the pulmonary veins, although the atrium may also be responsible for triggering or perpetuating atrial fibrillation. Ventricular tachycardia, often found in patients with ischemic heart disease and a history of myocardial infarction, is often life-threatening; it accounts for about 50% of sudden deaths. Treatment of Tachycardia The first line of treatment for tachycardia is antiarrhythmic drugs; for atrial fibrillation, anticoagulation drugs are also used to prevent stroke. For patients refractory to or unable to tolerate antiarrhythmic drugs, ablation of the arrhythmogenic heart tissues is the only option. Surgical ablation such as the Cox-Maze procedure is more invasive. Catheter ablation, involving the delivery of energy (most commonly radiofrequency) via a percutaneous catheter system guided by X-ray fluoroscopy, has been used in place of surgical ablation for many patients. However, this conventional approach in catheter ablation has not been found to be effective for the treatment of complex arrhythmias such as chronic atrial fibrillation or ventricular tachycardia. Advanced nonfluoroscopic mapping systems have been developed for guiding the ablation of these complex arrhythmias. Four nonfluoroscopic advanced mapping systems have been licensed by Health Canada: CARTO EP mapping System (manufactured by Biosense Webster, CA) uses weak magnetic fields and a special mapping/ablation catheter with a magnetic sensor to locate the catheter and reconstruct a 3-dimensional geometry of the heart superimposed with colour-coded electric potential maps to guide ablation. EnSite System (manufactured by Endocardial Solutions Inc., MN) includes a multi-electrode non-contact catheter that conducts simultaneous mapping. A processing unit uses the electrical data to computes more than 3,000 isopotential electrograms that are displayed on a reconstructed 3-dimensional geometry of the heart chamber. The navigational system, EnSite NavX, can be used separately with most mapping catheters. The LocaLisa Intracardiac System (manufactured by Medtronics Inc, MN) is a navigational system that uses an electrical field to locate the mapping catheter. It reconstructs the location of the electrodes on the mapping catheter in 3-dimensional virtual space, thereby enabling an ablation catheter to be directed to the electrode that identifies abnormal electric potential. Polar Constellation Advanced Mapping Catheter System (manufactured by Boston Scientific, MA) is a multielectrode basket catheter with 64 electrodes on 8 splines. Once deployed, each electrode is automatically traced. The information enables a 3-dimensional model of the basket catheter to be computed. Colour-coded activation maps are reconstructed online and displayed on a monitor. By using this catheter, a precise electrical map of the atrium can be obtained in several heartbeats. A systematic search of Cochrane, MEDLINE and EMBASE was conducted to identify studies that compared ablation guided by any of the advanced systems to fluoroscopy-guided ablation of tachycardia. English-language studies with sample sizes greater than or equal to 20 that were published between 2000 and 2005 were included. Observational studies on safety of advanced mapping systems and fluoroscopy were also included. Outcomes of interest were acute success, defined as termination of arrhythmia immediately following ablation; long-term success, defined as being arrhythmia free at follow-up; total procedure time; fluoroscopy time; radiation dose; number of radiofrequency pulses; complications; cost; and the cost-effectiveness ratio. Quality of the individual studies was assessed using established criteria. Quality of the overall evidence was determined by applying the GRADE evaluation system. (3) Qualitative synthesis of the data was performed. Quantitative analysis using Revman 4.2 was performed when appropriate. Quality of the Studies Thirty-four studies met the inclusion criteria. These comprised 18 studies on CARTO (4 randomized controlled trials [RCTs] and 14 non-RCTs), 3 RCTs on EnSite NavX, 4 studies on LocaLisa Navigational System (1 RCT and 3 non-RCTs), 2 studies on EnSite and CARTO, 1 on Polar Constellation basket catheter, and 7 studies on radiation safety. The quality of the studies ranged from moderate to low. Most of the studies had small sample sizes with selection bias, and there was no blinding of patients or care providers in any of the studies. Duration of follow-up ranged from 6 weeks to 29 months, with most having at least 6 months of follow-up. There was heterogeneity with respect to the approach to ablation, definition of success, and drug management before and after the ablation procedure. Evidence is based on a small number of small RCTS and non-RCTS with methodological flaws.Advanced nonfluoroscopy mapping/navigation systems provided real time 3-dimensional images with integration of anatomic and electrical potential information that enable better visualization of areas of interest for ablationAdvanced nonfluoroscopy mapping/navigation systems appear to be safe; they consistently shortened the fluoroscopy duration and radiation exposure.Evidence suggests that nonfluoroscopy mapping and navigation systems may be used as adjuncts to rather than replacements for fluoroscopy in guiding the ablation of complex arrhythmias.Most studies showed a nonsignificant trend toward lower overall failure rate for advanced mapping-guided ablation compared with fluoroscopy-guided mapping.Pooled analyses of small RCTs and non-RCTs that compared fluoroscopy- with nonfluoroscopy-guided ablation of atrial fibrillation and atrial flutter showed that advanced nonfluoroscopy mapping and navigational systems:Yielded acute success rates of 69% to 100%, not significantly different from fluoroscopy ablation.Had overall failure rates at 3 months to 19 months of 1% to 40% (median 25%).Resulted in a 10% relative reduction in overall failure rate for advanced mapping guided-ablation compared to fluoroscopy guided ablation for the treatment of atrial fibrillation.Yielded added benefit over fluoroscopy in guiding the ablation of complex arrhythmia. The advanced systems were shown to reduce the arrhythmia burden and the need for antiarrhythmic drugs in patients with complex arrhythmia who had failed fluoroscopy-guided ablationBased on predominantly observational studies, circumferential PV ablation guided by a nonfluoroscopy system was shown to do the following:Result in freedom from atrial fibrillation (with or without antiarrhythmic drug) in 75% to 95% of patients (median 79%). This effect was maintained up to 28 months.Result in freedom from atrial fibrillation without antiarrhythmic drugs in 47% to 95% of patients (median 63%).Improve patient survival at 28 months after the procedure as compared with drug therapy.Require special skills; patient outcomes are operator dependent, and there is a significant learning curve effect.Complication rates of pulmonary vein ablation guided by an advanced mapping/navigation system ranged from 0% to 10% with a median of 6% during a follow-up period of 6 months to 29 months.The complication rate of the study with the longest follow-up was 8%.The most common complications of advanced catheter-guided ablation were stroke, transient ischemic attack, cardiac tamponade, myocardial infarction, atrial flutter, congestive heart failure, and pulmonary vein stenosis. A small number of cases with fatal atrial-esophageal fistula had been reported and were attributed to the high radiofrequency energy used rather than to the advanced mapping systems. An Ontario-based economic analysis suggests that the cumulative incremental upfront costs of catheter ablation of atrial fibrillation guided by advanced nonfluoroscopy mapping could be recouped in 4.7 years through cost avoidance arising from less need for antiarrhythmic drugs and fewer hospitalization for stroke and heart failure. Expert Opinion Expert consultants to the Medical Advisory Secretariat noted the following: Nonfluoroscopy mapping is not necessary for simple ablation procedures (e.g., typical flutter). However, it is essential in the ablation of complex arrhythmias including these:Symptomatic, drug-refractory atrial fibrillationArrhythmias in people who have had surgery for congenital heart disease (e.g., macro re-entrant tachycardia in people who have had surgery for congenital heart disease).Ventricular tachycardia due to myocardial infarctionAtypical atrial flutterAdvanced mapping systems represent an enabling technology in the ablation of complex arrhythmias. The ablation of these complex cases would not have been feasible or advisable with fluoroscopy-guided ablation and, therefore, comparative studies would not be feasible or ethical in such cases. (ABSTRACT TRUNCATED)

  11. The use of computerized image guidance in lumbar disk arthroplasty.

    PubMed

    Smith, Harvey E; Vaccaro, Alexander R; Yuan, Philip S; Papadopoulos, Stephen; Sasso, Rick

    2006-02-01

    Surgical navigation systems have been increasingly studied and applied in the application of spinal instrumentation. Successful disk arthroplasty requires accurate midline and rotational positioning for optimal function and longevity. A surgical simulation study in human cadaver specimens was done to evaluate and compare the accuracy of standard fluoroscopy, computer-assisted fluoroscopic image guidance, and Iso-C3D image guidance in the placement of lumbar intervertebral disk replacements. Lumbar intervertebral disk prostheses were placed using three different image guidance techniques in three human cadaver spine specimens at multiple levels. Postinstrumentation accuracy was assessed with thin-cut computed tomography scans. Intervertebral disk replacements placed using the StealthStation with Iso-C3D were more accurately centered than those placed using the StealthStation with FluoroNav and standard fluoroscopy. Intervertebral disk replacements placed with Iso-C3D and FluoroNav had improved rotational divergence compared with standard fluoroscopy. Iso-C3D and FluoroNav had a smaller interprocedure variance than standard fluoroscopy. These results did not approach statistical significance. Relative to both virtual and standard fluoroscopy, use of the StealthStation with Iso-C3D resulted in improved accuracy in centering the lumbar disk prosthesis in the coronal midline. The StealthStation with FluoroNav appears to be at least equivalent to standard fluoroscopy and may offer improved accuracy with rotational alignment while minimizing radiation exposure to the surgeon. Surgical guidance systems may offer improved accuracy and less interprocedure variation in the placement of intervertebral disk replacements than standard fluoroscopy. Further study regarding surgical navigation systems for intervertebral disk replacement is warranted.

  12. Radiation Dose Reduction in CT Fluoroscopy-Guided Cervical Transforaminal Epidural Steroid Injection by Modifying Scout and Planning Steps

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Paik, Nam Chull, E-mail: pncspine@gmail.com

    Background and PurposeIn CT fluoroscopy (CTF)-guided cervical transforaminal epidural steroid injection (TFESI), the majority of radiation dose is contributed by the planning CT scan rather than the CTF procedure itself. We replaced the planning helical CT with a spot CTF and accordingly changed the patient posture during scout and planning scans. The aim of this study was to test whether radiation dose reduction would be achieved by this protocol modification while still maintaining technical performance.MethodsOverall, 338 consecutive procedures before (control group: n = 163) and after (study group: n = 175) instituting the above-mentioned protocol modification were analyzed retrospectively, comparing patient characteristics (age, sex,more » neck diameter, and level injected) and technical performance [technical success rate, dose-length product (DLP), inadvertent contrast flow incidence, number of CTF acquisitions, and procedural time] between the two groups.ResultsAll injections were technically successful at every level from C3–C4 to C7–T1 without serious complications in both groups. The median DLP of the study group (7.92 mGy·cm) was significantly reduced compared to that of the control group (39.05 mGy·cm, P < 0.001). There were no significant differences between the two groups regarding the incidence of inadvertent contrast flow (20.6 vs. 17.2 %, P = 0.426), number of CTF acquisitions (median 5 vs. 4, P = 0.123), and the procedural time (median 6.62 vs. 6.90 min, P = 0.100).ConclusionsWhen conducting CTF-guided cervical TFESIs, a significant radiation dose reduction (median 79.7 % in DLP) can be achieved by modifying scout and planning steps, without compromising the technical performance.« less

  13. Retrospective study of rapid-exchange monorail versus over-the-wire technique for femoropopliteal angioplasty.

    PubMed

    Jahnke, Thomas; Schäfer, Jost Philipp; Bolte, Hendrik; Schäfer, Fritz; Michalek, Jens; Charalambous, Nicholas; Sapoval, Marc; Müller-Hülsbeck, Stefan

    2008-01-01

    The purpose of this study was to compare procedural outcome of rapid-exchange (RX) monorail versus conventional over-the-wire (OTW) technique for femoropopliteal angioplasty. Demographic data, procedure details, angioplasty success, and complications of 328 consecutive percutaneous transluminal angioplasties (PTAs) were collected from a prospective database and retrospectively analyzed. Procedure details included duration of fluoroscopy, area-dose product, amount of contrast agent, sheath sizes, access route, length of stenosis, presence of total occlusion, technical and anatomical success (residual stenosis < 30% in the absence of complications), need for bail-out stenting, and periprocedural complications. The RX technique alone was used in 102 of 328 cases (31%); the OTW technique, in 226 of 328 of cases (68%). Technical success was 98% for the RX versus 95.4% for the OTW technique (p = 0.2). A significantly greater number of stents had to be implanted due to angioplasty failure when the OTW technique was used (RX, 5.9%; OTW, 13.7%; p = 0.04). There were no significant differences in fluoroscopy time, dose-area product, or amount of contrast medium used. The RX system facilitated the use of smaller sheath sizes (5 Fr = 38% and 6 Fr = 59% for RX versus 5 Fr = 16.8% and > or = 6 Fr = 82.5% for OTW) but showed only a tendency toward lower overall complication rates (16.6% [17/102] in the RX group versus 19.9% [45/226] in the OTW group; p = 0.09). There was no effect on length of hospitalization. RX monorail systems were not associated with higher procedural costs when compared to conventional OTW technique. We conclude that RX monorail systems seem to enhance the technical success of femoropopliteal angioplasty. Although smaller sheath sizes can be used due to the lower profile of the RX systems, there is only a tendency toward lower complication rates.

  14. Best Practices During Hip Arthroscopy: Aggregate Recommendations of High-Volume Surgeons.

    PubMed

    Gupta, Asheesh; Suarez-Ahedo, Carlos; Redmond, John M; Gerhardt, Michael B; Hanypsiak, Bryan; Stake, Christine E; Finch, Nathan A; Domb, Benjamin G

    2015-09-01

    To survey surgeons who perform a high volume of hip arthroscopy procedures regarding their operative technique, type of procedure, and postoperative management. We conducted a cross-sectional survey of 27 high-volume orthopaedic surgeons specializing in hip arthroscopy to report their preferences and practices related to their operative practice and postoperative rehabilitation protocol. All participants completed the survey in person in an anonymous fashion during a meeting of the American Hip Institute. All surgeons perform hip arthroscopy with the patient in the supine position, accessing the central compartment of the hip initially, using intraoperative fluoroscopy. All surgeons perform labral repair (100%), with the majority performing labral reconstructions (77.8%) and gluteus medius repairs (81.5%). There is variability in the type of anchors used during labral repair. Most surgeons perform capsular closure in most cases (88.9%), inject either intra-articular cortisone or platelet-rich plasma at the conclusion of the procedure (59%), and prescribe a postoperative hip brace for some or all patients (59%). There is considerable variability in rehabilitation protocols. All surgeons routinely prescribe postoperative heterotopic ossification prophylaxis to their patients, with most surgeons (88.9%) prescribing a nonsteroidal anti-inflammatory medication for 3 weeks. Forty percent of the respondents use the modified Harris Hip Score as the most important outcome measure. Consistent practices such as use of intraoperative fluoroscopy, heterotopic ossification prophylaxis, and labral repair skills were identified by surveying 27 hip arthroscopy surgeons at high-volume centers. Most of the surgeons performed routine capsular closure unless underlying conditions precluded capsular release or plication. The survey identified higher variability between surgeons regarding postoperative rehabilitation protocols and use of intra-articular pharmacologic injections at the end of the procedure. These data may provide surgeons with a set of aggregate trends that may help guide training, clinical practice, and research in the evolving field of hip arthroscopy. Copyright © 2015 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

  15. Visuospatial and psychomotor aptitude predicts endovascular performance of inexperienced individuals on a virtual reality simulator.

    PubMed

    Van Herzeele, Isabelle; O'Donoghue, Kevin G L; Aggarwal, Rajesh; Vermassen, Frank; Darzi, Ara; Cheshire, Nicholas J W

    2010-04-01

    This study evaluated virtual reality (VR) simulation for endovascular training of medical students to determine whether innate perceptual, visuospatial, and psychomotor aptitude (VSA) can predict initial and plateau phase of technical endovascular skills acquisition. Twenty medical students received didactic and endovascular training on a commercially available VR simulator. Each student treated a series of 10 identical noncomplex renal artery stenoses endovascularly. The simulator recorded performance data instantly and objectively. An experienced interventionalist rated the performance at the initial and final sessions using generic (out of 40) and procedure-specific (out of 30) rating scales. VSA were tested with fine motor dexterity (FMD, Perdue Pegboard), psychomotor ability (minimally invasive virtual reality surgical trainer [MIST-VR]), image recall (Rey-Osterrieth), and organizational aptitude (map-planning). VSA performance scores were correlated with the assessment parameters of endovascular skills at commencement and completion of training. Medical students exhibited statistically significant learning curves from the initial to the plateau performance for contrast usage (medians, 28 vs 17 mL, P < .001), total procedure time (2120 vs 867 seconds, P < .001), and fluoroscopy time (993 vs. 507 seconds, P < .001). Scores on generic and procedure-specific rating scales improved significantly (10 vs 25, P < .001; 8 vs 17 P < .001). Significant correlations were noted for FMD with initial and plateau sessions for fluoroscopy time (r(s) = -0.564, P = .010; r(s) = -.449, P = .047). FMD correlated with procedure-specific scores at the initial session (r(s) = .607, P = .006). Image recall correlated with generic skills at the end of training (r(s) = .587, P = .006). Simulator-based training in endovascular skills improved performance in medical students. There were significant correlations between initial endovascular skill and fine motor dexterity as well as with image recall at end of the training period. In addition to current recruitment strategies, VSA may be a useful tool for predictive validity studies.

  16. Shoulder-Mounted Robot for MRI-guided arthrography: Accuracy and mounting study.

    PubMed

    Monfaredi, R; Wilson, E; Sze, R; Sharma, K; Azizi, B; Iordachita, I; Cleary, K

    2015-08-01

    A new version of our compact and lightweight patient-mounted MRI-compatible 4 degree-of-freedom (DOF) robot for MRI-guided arthrography procedures is introduced. This robot could convert the traditional two-stage arthrography procedure (fluoroscopy-guided needle insertion followed by a diagnostic MRI scan) to a one-stage procedure, all in the MRI suite. The results of a recent accuracy study are reported. A new mounting technique is proposed and the mounting stability is investigated using optical and electromagnetic tracking on an anthropomorphic phantom. Five volunteer subjects including 2 radiologists were asked to conduct needle insertion in 4 different random positions and orientations within the robot's workspace and the displacement of the base of the robot was investigated during robot motion and needle insertion. Experimental results show that the proposed mounting method is stable and promising for clinical application.

  17. Can We Improve Workflows in the OR? A Comparison of Quality Perceptions and Preoperative Efficiency across Institutions in Spine Surgery.

    PubMed

    Wasterlain, Amy S; Tran, Andrew A; Tang, Chad; Campbell, David R; Braun, Hillary J; Scuderi, Yasmeen A; Scuderi, Gaetano J

    2015-03-01

    Cost containment and surgical inefficiencies are major concerns for hospitals in this era of declining resources. The primary aim of this investigation was to understand subjective perceptions of perioperative spine surgical quality across three practice settings and to identify potential factors contributing to these perceptions. Subsequently, we objectively evaluated factors that influence the duration of time in which the patient is in the operating room (OR) prior to the surgical incision and assessed the influence of fluoroscopy technician expertise on radiation dose and imaging efficiency. One hundred and eight medical device representatives with at least 1 year of OR experience were surveyed at a national conference. Three distinct healthcare facilities were identified: university, small volume, and large volume private hospitals. Respondents rated facilities on a five-point scale for staff quality; size and consistency of surgical teams; and overall likelihood of recommending the facility. Separately, 140 posterior lumbar procedures from two institutions were retrospectively reviewed. Two time periods were quantified for each surgical case: patient arrival in the OR to induction of anesthesia (T1) and induction to surgical incision (T2). T1 and T2 were compared between university and large private hospital settings using t tests and multivariate analysis. For 44 separate lumbar spine surgical procedures, practice setting, patient BMI, number of vertebral levels requiring imaging, number of localizing fluoroscopy images taken, total fluoroscopy time, total radiation dose, fluoroscopy machine, and whether the fluoroscopist could correctly state his or her role, which was to obtain a lateral lumbar localizing image, were recorded. T-tests were used to compare cases in which the fluoroscopist could and could not correctly state the task. Survey ratings for surgeons were not significantly different across university, large private, and small private hospitals. Fewer circulating nurses were rated as excellent or good in university versus private hospitals (p < 0.001). Small volume private hospital surgical teams were more likely to have worked together before than university teams (p < 0.05), and university teams were larger (p < 0.05). Respondents were more likely to recommend a university or large private hospital for complex instrumentation cases (p < 0.001). On objective measures, university patients were older, less obese, and had higher mean ASA scores (2.5 versus 2.2, p < 0.001). Compared to the university setting, private hospital cases had significantly shorter Time 1 (8 versus 37 min, p < 0.001) and Time 2 (23 versus 30 min, p < 0.001), even after adjusting for ASA score, BMI, and age. Cases in which the fluoroscopist knew the imaging purpose were associated with significantly fewer images (mean 1.8 versus 3.4 images, p < 0.0001) and shorter total exposure times (2.3 versus 4.0 sec, p < 0.001). Operations performed in the university setting were associated with significantly more images (2.7 versus 1.8 images, p < 0.001), longer total exposure times (3.2 versus 2.3 sec, p = 0.0027), and total radiation dose (27.8 versus 53.3 rad, p < 0.001) when compared with those performed in the private setting. The university practice setting was associated with significantly more images (2.7 versus 1.8 images, p < 0.001), longer total exposure times (3.2 versus 2.3 sec, p = 0.003), and total radiation dose (27.8 versus 53.3 rad, p < 0.001) when compared with non-university settings. Large private and university hospitals had higher surgeon ratings. The university setting was associated with larger and less consistent surgical teams and lower nurse ratings. Surgical staff awareness of the procedure and attention to preoperative tasks specific to the procedure reduced pre-operative time spent in the OR as well as fluoroscopy radiation. These data suggest that nurses and support staff make substantial contributions to overall quality of care, and that leadership and interpersonal coordination are especially important within large teams at teaching hospitals.

  18. An Assessment of Common Approaches to Estimating Peak Skin Dose Resulting From Fluoroscopically Guided Interventions

    NASA Astrophysics Data System (ADS)

    Smith, Caleb Martin

    Fluoroscopy guided procedures are increasing in complexity, and with that, Peak Skin Doses (PSD) that produce cutaneous radiation injury are a growing concern. Direct measurement of PSD is possible, but the decision to do so must be made in advance. PSD estimates and correctly monitoring their possible deterministic skin injuries are important to patient care. Three methods of indirect PSD estimation are examined for nine cases at MedStar Georgetown University Hospital. The aim of the study is to determine the magnitude of variation between these three methods for estimating the PSD. Method 1 (Fluoroscopy Time and Maximum Entrance Skin Exposure) was used at MedStar Georgetown University Hospital up until 2016. Methods 2 and 3 incorporate procedure information (Reference Point Air Kerma, Source-to-Patent distance, and Backscatter Factor) from DICOM (Digital Imaging and Communications in Medicine) tags into PSD estimates. Method 1 PSD estimates are vastly different, by as much as 136%, than those from Methods 2 and 3. Method 2 and 3 PSD estimates differ very little, 7.3% or less. Governing bodies have discounted Method 1 as a reliable dose metric because of its poor correlation with PSD. The accuracy of Method 2 is suitable to determine PSD and which dose band a patient fits so their injuries can be accurately monitored. Method 3, the most time intensive approach, should only be used in the case of a sentinel event where a full investigation is warranted.

  19. Risk of whole body radiation exposure and protective measures in fluoroscopically guided interventional techniques: a prospective evaluation.

    PubMed

    Manchikanti, Laxmaiah; Cash, Kim A; Moss, Tammy L; Rivera, Jose; Pampati, Vidyasagar

    2003-08-06

    BACKGROUND: Fluoroscopic guidance is frequently utilized in interventional pain management. The major purpose of fluoroscopy is correct needle placement to ensure target specificity and accurate delivery of the injectate. Radiation exposure may be associated with risks to physician, patient and personnel. While there have been many studies evaluating the risk of radiation exposure and techniques to reduce this risk in the upper part of the body, the literature is scant in evaluating the risk of radiation exposure in the lower part of the body. METHODS: Radiation exposure risk to the physician was evaluated in 1156 patients undergoing interventional procedures under fluoroscopy by 3 physicians. Monitoring of scattered radiation exposure in the upper and lower body, inside and outside the lead apron was carried out. RESULTS: The average exposure per procedure was 12.0 PlusMinus; 9.8 seconds, 9.0 PlusMinus; 0.37 seconds, and 7.5 PlusMinus; 1.27 seconds in Groups I, II, and III respectively. Scatter radiation exposure ranged from a low of 3.7 PlusMinus; 0.29 seconds for caudal/interlaminar epidurals to 61.0 PlusMinus; 9.0 seconds for discography. Inside the apron, over the thyroid collar on the neck, the scatter radiation exposure was 68 mREM in Group I consisting of 201 patients who had a total of 330 procedures with an average of 0.2060 mREM per procedure and 25 mREM in Group II consisting of 446 patients who had a total of 662 procedures with average of 0.0378 mREM per procedure. The scatter radiation exposure was 0 mREM in Group III consisting of 509 patients who had a total 827 procedures. Increased levels of exposures were observed in Groups I and II compared to Group III, and Group I compared to Group II.Groin exposure showed 0 mREM exposure in Groups I and II and 15 mREM in Group III. Scatter radiation exposure for groin outside the apron in Group I was 1260 mREM and per procedure was 3.8182 mREM. In Group II the scatter radiation exposure was 400 mREM and with 0.6042 mREM per procedure. In Group III the scatter radiation exposure was 1152 mREM with 1.3930 mREM per procedure. CONCLUSION: Results of this study showed that scatter radiation exposure to both the upper and lower parts of the physician's body is present. Protection was offered by traditional measures to the upper body only.

  20. Analysis of Patients' X-ray Exposure in 146 Percutaneous Radiologic Gastrostomies.

    PubMed

    Petersen, Tim-Ole; Reinhardt, Martin; Fuchs, Jochen; Gosch, Dieter; Surov, Alexey; Stumpp, Patrick; Kahn, Thomas; Moche, Michael

    2017-09-01

    Purpose  Analysis of patient´s X-ray exposure during percutaneous radiologic gastrostomies (PRG) in a larger population. Materials and Methods  Data of primary successful PRG-procedures, performed between 2004 and 2015 in 146 patients, were analyzed regarding the exposition to X-ray. Dose-area-product (DAP), dose-length-product (DLP) respectively, and fluoroscopy time (FT) were correlated with the used x-ray systems (Flatpanel Detector (FD) vs. Image Itensifier (BV)) and the necessity for periprocedural placement of a nasogastric tube. Additionally, the effective X-ray dose for PRG placement using fluoroscopy (DL), computed tomography (CT), and cone beam CT (CBCT) was estimated using a conversion factor. Results  The median DFP of PRG-placements under fluoroscopy was 163 cGy*cm 2 (flat panel detector systems: 155 cGy*cm 2 ; X-ray image intensifier: 175 cGy*cm 2 ). The median DLZ was 2.2 min. Intraprocedural placement of a naso- or orogastric probe (n = 68) resulted in a significant prolongation of the median DLZ to 2.5 min versus 2 min in patients with an already existing probe. In addition, dose values were analyzed in smaller samples of patients in which the PRG was placed under CBCT (n = 7, median DFP = 2635 cGy*cm 2 ), or using CT (n = 4, median DLP = 657 mGy*cm). Estimates of the median DFP and DLP showed effective doses of 0.3 mSv for DL-assisted placements (flat panel detector 0.3 mSv, X-ray image converter 0.4 mSv), 7.9 mSv using a CBCT - flat detector, and 9.9 mSv using CT. This corresponds to a factor 26 of DL versus CBCT, or a factor 33 of DL versus CT. Conclusion  In order to minimize X-ray exposure during PRG-procedures for patients and staff, fluoroscopically-guided interventions should employ flat detector systems with short transmittance sequences in low dose mode and with slow image frequency. Series recordings can be dispensed with. The intraprocedural placement of a naso- or orogastric probe significantly extends FT, but has little effect on the overall dose of the intervention. Due to the significantly higher X-ray exposure, the use of a CBCT as well as PRG-placements using CT should be limited to clinically absolutely necessary exceptions with strict indication. Key Points   · Fluoroscopically-guided PRG placements are interventions with low X-ray exposure.. · X-ray exposure from fluoroscopy is lower using flat panel detector systems as compared to image intensifier systems.. · The concomitant placement of an oro- or nasogastric probe extends the fluoroscopy time.. · Gastric probe placement is worthwhile to prevent the premature use of the significantly radiation-intensive CT.. · The use of the C-arm CT or the CT increases the beam exposure by 26 or 33 times, respectively.. · The PRG placement using C-arm CT and CT should only be performed in exceptional cases.. Citation Format · Petersen TO, Reinhardt M, Fuchs J et al. Analysis of Patients' X-ray Exposure in 146 Percutaneous Radiologic Gastrostomies. Fortschr Röntgenstr 2017; 189: 820 - 827. © Georg Thieme Verlag KG Stuttgart · New York.

  1. Fluoroscopy

    MedlinePlus

    ... through a clinical facility’s quality assurance program, are fundamental to radiation protection. More information about the principles ... as part of quality assurance program emphasizing radiation management. Health care providers who use fluoroscopy should be ...

  2. Simulation-based educational curriculum for fluoroscopically guided lumbar puncture improves operator confidence and reduces patient dose.

    PubMed

    Faulkner, Austin R; Bourgeois, Austin C; Bradley, Yong C; Hudson, Kathleen B; Heidel, R Eric; Pasciak, Alexander S

    2015-05-01

    Fluoroscopically guided lumbar puncture (FGLP) is a commonly performed procedure with increased success rates relative to bedside technique. However, FGLP also exposes both patient and staff to ionizing radiation. The purpose of this study was to determine if the use of a simulation-based FGLP training program using an original, inexpensive lumbar spine phantom could improve operator confidence and efficiency, while also reducing patient dose. A didactic and simulation-based FGLP curriculum was designed, including a 1-hour lecture and hands-on training with a lumbar spine phantom prototype developed at our institution. Six incoming post-graduate year 2 (PGY-2) radiology residents completed a short survey before taking the course, and each resident practiced 20 simulated FGLPs using the phantom before their first clinical procedure. Data from the 114 lumbar punctures (LPs) performed by the six trained residents (prospective cohort) were compared to data from 514 LPs performed by 17 residents who did not receive simulation-based training (retrospective cohort). Fluoroscopy time (FT), FGLP success rate, and indication were compared. There was a statistically significant reduction in average FT for the 114 procedures performed by the prospective study cohort compared to the 514 procedures performed by the retrospective cohort. This held true for all procedures in aggregate, LPs for myelography, and all procedures performed for a diagnostic indication. Aggregate FT for the prospective group (0.87 ± 0.68 minutes) was significantly lower compared to the retrospective group (1.09 ± 0.65 minutes) and resulted in a 25% reduction in average FT (P = .002). There was no statistically significant difference in the number of failed FGLPs between the two groups. Our simulation-based FGLP curriculum resulted in improved operator confidence and reduced FT. These changes suggest that resident procedure efficiency was improved, whereas patient dose was reduced. The FGLP training program was implemented by radiology residents and required a minimal investment of time and resources. The LP spine phantom used during training was inexpensive, durable, and effective. In addition, the phantom is compatible with multiple modalities including fluoroscopy, computed tomography, and ultrasound and could be easily adapted to other applications such as facet injections or joint arthrograms. Copyright © 2015 AUR. Published by Elsevier Inc. All rights reserved.

  3. Occupational Radiation Exposure During Endovascular Aortic Repair

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sailer, Anna M., E-mail: anni.sailer@mumc.nl; Schurink, Geert Willem H., E-mail: gwh.schurink@mumc.nl; Bol, Martine E., E-mail: m.bol@maastrichtuniversity.nl

    PurposeThe aim of the study was to evaluate the radiation exposure to operating room personnel and to assess determinants for high personal doses during endovascular aortic repair.Materials and MethodsOccupational radiation exposure was prospectively evaluated during 22 infra-renal aortic repair procedures (EVAR), 11 thoracic aortic repair procedures (TEVAR), and 11 fenestrated or branched aortic repair procedures (FEVAR). Real-time over-lead dosimeters attached to the left breast pocket measured personal doses for the first operators (FO) and second operators (SO), radiology technicians (RT), scrub nurses (SN), anesthesiologists (AN), and non-sterile nurses (NSN). Besides protective apron and thyroid collar, no additional radiation shielding wasmore » used. Procedural dose area product (DAP), iodinated contrast volume, fluoroscopy time, patient’s body weight, and C-arm angulation were documented.ResultsAverage procedural FO dose was significantly higher during FEVAR (0.34 ± 0.28 mSv) compared to EVAR (0.11 ± 0.21 mSv) and TEVAR (0.06 ± 0.05 mSv; p = 0.003). Average personnel doses were 0.17 ± 0.21 mSv (FO), 0.042 ± 0.045 mSv (SO), 0.019 ± 0.042 mSv (RT), 0.017 ± 0.031 mSv (SN), 0.006 ± 0.007 mSv (AN), and 0.004 ± 0.009 mSv (NSN). SO and AN doses were strongly correlated with FO dose (p = 0.003 and p < 0.001). There was a significant correlation between FO dose and procedural DAP (R = 0.69, p < 0.001), iodinated contrast volume (R = 0.67, p < 0.001) and left-anterior C-arm projections >60° (p = 0.02), and a weak correlation with fluoroscopy time (R = 0.40, p = 0.049).ConclusionAverage FO dose was a factor four higher than SO dose. Predictors for high personal doses are procedural DAP, iodinated contrast volume, and left-anterior C-arm projections greater than 60°.« less

  4. Architecture of a high-performance surgical guidance system based on C-arm cone-beam CT: software platform for technical integration and clinical translation

    NASA Astrophysics Data System (ADS)

    Uneri, Ali; Schafer, Sebastian; Mirota, Daniel; Nithiananthan, Sajendra; Otake, Yoshito; Reaungamornrat, Sureerat; Yoo, Jongheun; Stayman, J. Webster; Reh, Douglas; Gallia, Gary L.; Khanna, A. Jay; Hager, Gregory; Taylor, Russell H.; Kleinszig, Gerhard; Siewerdsen, Jeffrey H.

    2011-03-01

    Intraoperative imaging modalities are becoming more prevalent in recent years, and the need for integration of these modalities with surgical guidance is rising, creating new possibilities as well as challenges. In the context of such emerging technologies and new clinical applications, a software architecture for cone-beam CT (CBCT) guided surgery has been developed with emphasis on binding open-source surgical navigation libraries and integrating intraoperative CBCT with novel, application-specific registration and guidance technologies. The architecture design is focused on accelerating translation of task-specific technical development in a wide range of applications, including orthopaedic, head-and-neck, and thoracic surgeries. The surgical guidance system is interfaced with a prototype mobile C-arm for high-quality CBCT and through a modular software architecture, integration of different tools and devices consistent with surgical workflow in each of these applications is realized. Specific modules are developed according to the surgical task, such as: 3D-3D rigid or deformable registration of preoperative images, surgical planning data, and up-to-date CBCT images; 3D-2D registration of planning and image data in real-time fluoroscopy and/or digitally reconstructed radiographs (DRRs); compatibility with infrared, electromagnetic, and video-based trackers used individually or in hybrid arrangements; augmented overlay of image and planning data in endoscopic or in-room video; real-time "virtual fluoroscopy" computed from GPU-accelerated DRRs; and multi-modality image display. The platform aims to minimize offline data processing by exposing quantitative tools that analyze and communicate factors of geometric precision. The system was translated to preclinical phantom and cadaver studies for assessment of fiducial (FRE) and target registration error (TRE) showing sub-mm accuracy in targeting and video overlay within intraoperative CBCT. The work culminates in the development of a CBCT guidance system (reported here for the first time) that leverages the technical developments in Carm CBCT and associated technologies for realizing a high-performance system for translation to clinical studies.

  5. WE-EF-303-05: Development and Commissioning of Real-Time Imaging Function for Respiratory-Gated Spot-Scanning Proton Beam Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Miyamoto, N; Takao, S; Matsuura, T

    2015-06-15

    Purpose: To realize real-time-image gated proton beam therapy (RGPT) for treating mobile tumors. Methods: The rotating gantry of spot scanning proton beam therapy has been designed to equip two x-ray fluoroscopy devices that enable real-time imaging of the internal fiducial markers during respiration. Three-dimensional position of the fiducial marker located near the tumor can be calculated from the fluoroscopic images obtained from orthogonal directions and therapeutic beam is gated only when the fiducial marker is within the predefined gating window. Image acquisition rate can be selected from discrete value ranging from 0.1 Hz to 30 Hz. In order to confirmmore » the effectiveness of RGPT and apply it clinically, clinical commissioning was conducted. Commissioning tests were categorized to main three parts including geometric accuracy, temporal accuracy and dosimetric evaluation. Results: Developed real-time imaging function has been installed and its basic performances have been confirmed. In the evaluation of geometric accuracy, coincidence of three-dimensional treatment room coordinate system and imaging coordinate system was confirmed to be less than 1 mm. Fiducial markers (gold sphere and coil) were able to be tracked in simulated clinical condition using an anthropomorphic chest phantom. In the evaluation of temporal accuracy, latency from image acquisition to gate on/off signal was about 60 msec in typical case. In dosimetric evaluation, treatment beam characteristics including beam irradiation position and dose output were stable in gated irradiation. Homogeneity indices to the mobile target were 0.99 (static), 0.89 (w/o gating, motion is parallel to direction of scan), 0.75 (w/o gating, perpendicular), 0.98 (w/ gating, parallel) and 0.93 (w/ gating, perpendicular). Dose homogeneity to the mobile target can be maintained in RGPT. Conclusion: Real-time imaging function utilizing x-ray fluoroscopy has been developed and commissioned successfully in order to realize RGPT. Funding Support: This research was partially supported by Japan Society for the Promotion of Science (JSPS) through the FIRST Program. Conflict of Interest: Prof. Shirato has research fund from Hitachi Ltd, Mitsubishi Heavy Industries Ltd and Shimadzu Corporation.« less

  6. Fluoroscopy-Guided Percutaneous Vertebral Body Biopsy Using a Novel Drill-Powered Device: Technical Case Series

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wallace, Adam N., E-mail: wallacea@mir.wustl.edu; Pacheco, Rafael A., E-mail: pachecor@mir.wustl.edu; Tomasian, Anderanik, E-mail: tomasiana@mir.wustl.edu

    2016-02-15

    BackgroundA novel coaxial biopsy system powered by a handheld drill has recently been introduced for percutaneous bone biopsy. This technical note describes our initial experience performing fluoroscopy-guided vertebral body biopsies with this system, compares the yield of drill-assisted biopsy specimens with those obtained using a manual technique, and assesses the histologic adequacy of specimens obtained with drill assistance.MethodsMedical records of all single-level, fluoroscopy-guided vertebral body biopsies were reviewed. Procedural complications were documented according to the Society of Interventional Radiology classification. The total length of bone core obtained from drill-assisted biopsies was compared with that of matched manual biopsies. Pathology reportsmore » were reviewed to determine the histologic adequacy of specimens obtained with drill assistance.ResultsTwenty eight drill-assisted percutaneous vertebral body biopsies met study inclusion criteria. No acute complications were reported. Of the 86 % (24/28) of patients with clinical follow-up, no delayed complications were reported (median follow-up, 28 weeks; range 5–115 weeks). The median total length of bone core obtained from drill-assisted biopsies was 28 mm (range 8–120 mm). This was longer than that obtained from manual biopsies (median, 20 mm; range 5–45 mm; P = 0.03). Crush artifact was present in 11 % (3/28) of drill-assisted biopsy specimens, which in one case (3.6 %; 1/28) precluded definitive diagnosis.ConclusionsA drill-assisted, coaxial biopsy system can be used to safely obtain vertebral body core specimens under fluoroscopic guidance. The higher bone core yield obtained with drill assistance may be offset by the presence of crush artifact.« less

  7. Outcomes and Direct Costs of Inferior Vena Cava Filter Placement and Retrieval within the IR and Surgical Settings.

    PubMed

    Makary, Mina S; Kapke, Jordan; Yildiz, Vedat; Pan, Xueliang; Dowell, Joshua D

    2018-02-01

    To compare the outcomes and costs of inferior vena cava (IVC) filter placement and retrieval in the interventional radiology (IR) and surgical departments at a tertiary-care center. Retrospective review was performed of 142 sequential outpatient IVC filter placements and 244 retrievals performed in the IR suite and operating room (OR) from 2013 to 2016. Patient demographic data, procedural characteristics, outcomes, and direct costs were compared between cohorts. Technical success rates of 100% were achieved for both IR and OR filter placements, and 98% of filters were successfully retrieved by IR means, compared with 83% in the OR (P < .01). Fluoroscopy time was similar for IR and OR filter insertions, but IR retrievals required half the fluoroscopy time, with an average of 9 minutes vs 18 minutes in the OR (P = .02). There was no significant difference between cohorts in the incidences of complications for filter retrievals, but more postprocedural complications were observed for OR placements (8%) vs IR placements (1%; P = .05). The most severe complication occurred during an OR filter retrieval, resulting in entanglement of the snare device and conversion to an emergent open filter removal by vascular surgery. Direct costs were approximately 20% higher for OR vs IR IVC filter placements ($2,246 vs $2,671; P = .01). Filter placements are equally successfully performed in IR and OR settings, but OR patients experienced significantly higher postprocedural complication rates and incurred higher costs. In contrast, higher technical success rates and shorter fluoroscopy times were observed for IR filter retrievals compared with those performed in the OR. Copyright © 2017 SIR. Published by Elsevier Inc. All rights reserved.

  8. An unusual percutaneous transmitral commissurotomy: A collection of four rare occurrences!

    PubMed

    Shankarappa, Ravindranath K; Agrawal, Navin; Patra, Soumya; Karur, Satish; Nanjappa, Manjunath C

    2013-09-01

    We are presenting an interesting case of a 30-year-old patient taken for percutaneous transvenous mitral commissurotomy (PTMC) for severe rheumatic mitral stenosis in which there was a collection of four unusual occurrences during the course of a procedure. She had recurrent generalized tonic-clonic seizures immediately after femoral sheath insertion requiring the patient to be mechanically ventilated. Subsequently, the pressure tracings recorded with catheters in the aorta and the pulmonary artery showed transient unusually high supra-systemic pulmonary artery pressure. During inflation the Accura PTMC balloon which was used to dilate the mitral valve ruptured and the procedure subsequently had to be completed using another balloon catheter. During the procedure the presence of a distended stomach due to insufflations of air during positive pressure ventilation which subsided subsequently was another unusual documentation on fluoroscopy. The final outcome of the procedure was successful. This case presents an interesting collection of unusual occurrences during a PTMC procedure which started on an unusual note but ended on a successful one. Careful assessment and appropriate management of complications can lead to successful outcome of procedures as in our case.

  9. Interventional Cardiology: What's New?

    PubMed

    Scansen, Brian A

    2017-09-01

    Interventional cardiology in veterinary medicine continues to expand beyond the standard 3 procedures of patent ductus arteriosus occlusion, balloon pulmonary valvuloplasty, and transvenous pacing. Opportunities for fellowship training; advances in equipment, including high-resolution digital fluoroscopy, real-time 3-dimensional transesophageal echocardiography, fusion imaging, and rotational angiography; ultrasound-guided access and vascular closure devices; and refinement of techniques, including cutting and high-pressure ballooning, intracardiac and intravascular stent implantation, septal defect occlusion, transcatheter valve implantation, and hybrid approaches, are likely to transform the field over the next decade. Copyright © 2017 Elsevier Inc. All rights reserved.

  10. Occupational radiation doses during interventional procedures

    NASA Astrophysics Data System (ADS)

    Nuraeni, N.; Hiswara, E.; Kartikasari, D.; Waris, A.; Haryanto, F.

    2016-03-01

    Digital subtraction angiography (DSA) is a type of fluoroscopy technique used in interventional radiology to clearly visualize blood vessels in a bony or dense soft tissue environment. The use of DSA procedures has been increased quite significantly in the Radiology departments in various cities in Indonesia. Various reports showed that both patients and medical staff received a noticeable radiation dose during the course of this procedure. A study had been carried out to measure these doses among interventionalist, nurse and radiographer. The results show that the interventionalist and the nurse, who stood quite close to the X-ray beams compared with the radiographer, received radiation higher than the others. The results also showed that the radiation dose received by medical staff were var depending upon the duration and their position against the X-ray beams. Compared tothe dose limits, however, the radiation dose received by all these three medical staff were still lower than the limits.

  11. Eye dose to staff involved in interventional and procedural fluoroscopy

    NASA Astrophysics Data System (ADS)

    McLean, D.; Hadaya, D.; Tse, J.

    2016-03-01

    In 2011 the International Commission on Radiological Protection (ICRP) lowered the occupational eye dose limit from 150 to 20 mSv/yr [1]. While international jurisdictions are in a process of adopting these substantial changes, medical physicists at the clinical level have been advising medical colleagues on specific situations based on dose measurements. Commissioned and calibrated TLDs mounted in commercially available holders designed to simulate the measurement of Hp(3), were applied to staff involved in x-ray procedures for a one month period. During this period clinical procedure data was concurrently collected and subject to audit. The use or not of eye personal protective equipment (PPE) was noted for all staff. Audits were conducted in the cardiac catheterisation laboratory, the interventional angiography rooms and the procedural room where endoscopic retrograde cholangiopancreatography (ERCP) procedures are performed. Significant levels of occupational dose were recorded in the cardiac and interventional procedures, with maximum reading exceeding the new limit for some interventional radiologists. No significant eye doses were measured for staff performing ERCP procedures. One outcome of the studies was increased use of eye PPE for operators of interventional equipment with increased availability also to nursing staff, when standing in close proximity to the patient during procedures.

  12. Use of the Uro Dyna-CT in endourology – the new frontier

    PubMed Central

    Vicentini, Fabio C.; Botelho, Luiz A. A.; Braz, José L. M.; Almeida, Ernane de S.; Hisano, Marcelo

    2017-01-01

    ABSTRACT We describe the use of the Uro Dyna-CT, an imaging system used in the operating room that produces real-time three-dimensional (3D) imaging and cross-sectional image reconstructions similar to an intraoperative computerized tomography, during a percutaneous nephrolithotomy and a contralateral flexible ureteroscopy in a complete supine position. A 65 year-old female patient had an incomplete calyceal staghorn stone in the right kidney and a 10mm in the left one. The procedure was uneventful and the intraoperative use of the Uro Dyna-CT identified 2 residual stones that were not found by digital fluoroscopy and flexible nephroscopy at the end of surgery, helping us to render the patient stone-free in one procedure, which was confirmed by a postoperative CT scan. Prospective studies will define the real role of the Uro Dyna-CT for endourological procedures, but its use seems to be a very promising tool for improving stone free rates and decreasing auxiliary procedures, especially for complex cases. PMID:28338302

  13. Estimating Effective Dose from Phantom Dose Measurements in Atrial Fibrillation Ablation Procedures and Comparison of MOSFET and TLD Detectors in a Small Animal Dosimetry Setting

    NASA Astrophysics Data System (ADS)

    Anderson-Evans, Colin David

    Two different studies will be presented in this work. The first involves the calculation of effective dose from a phantom study which simulates an atrial fibrillation (AF) ablation procedure. The second involves the validation of metal-oxide semiconducting field effect transistors (MOSFET) for small animal dosimetry applications as well as improved characterization of the animal irradiators on Duke University's campus. Atrial Fibrillation is an ever increasing health risk in the United States. The most common type of cardiac arrhythmia, AF is associated with increased mortality and ischemic cerebrovascular events. Managing AF can include, among other treatments, an interventional procedure called catheter ablation. The procedure involves the use of biplane fluoroscopy during which a patient can be exposed to radiation for as much as two hours or more. The deleterious effects of radiation become a concern when dealing with long fluoroscopy times, and because the AF ablation procedure is elective, it makes relating the risks of radiation ever more essential. This study hopes to quantify the risk through the derivation of dose conversion coefficients (DCCs) from the dose-area product (DAP) with the intent that DCCs can be used to provide estimates of effective dose (ED) for typical AF ablation procedures. A bi-plane fluoroscopic and angiographic system was used for the simulated AF ablation procedures. For acquisition of organ dose measurements, 20 diagnostic MOSFET detectors were placed at selected organs in a male anthropomorphic phantom, and these detectors were attached to 4 bias supplies to obtain organ dose readings. The DAP was recorded from the system console and independently validated with an ionization chamber and radiochromic film. Bi-plane fluoroscopy was performed on the phantom for 10 minutes to acquire the dose rate for each organ, and the average clinical procedure time was multiplied by each organ dose rate to obtain individual organ doses. The effective dose was computed by summing the product of each organ dose and the corresponding tissue weighting factor from the ICRP publication 103. Further risk calculations were done according to the BEIR VII Phase 2 report to obtain relative and lifetime attributable risks of cancer for an average AF ablation procedure. The ED was computed separately for the biplane fluoroscopic and angiographic system's 'low' and 'normal fluoro' automated settings, yielding 27.9 mSv and 45.6 mSv respectively for an average procedure time of 88.2 minutes. The corresponding DAP was 48.7 Gy cm2 and 79.1 Gy cm2 for low and normal settings respectively. The independently measured DAP was found to be within 0.1 % of that measured by the fluoroscopy system's onboard flat panel detectors. DCCs were calculated to be 0.573 and 0.577 for the respective low and normal settings. The results proved to be very closely matched, which was to be expected. The calculated cancer risks were fairly low due to the age of most patients (less than 5 incidences of solid tumor per 100,000 exposed for liver colon and stomach; 100-300 incidences per 100,000 exposed for lungs), but concern remains that longer procedures could increase the risk of erythema or other serious skin injuries. The second section of this thesis study involves the quantification and distribution of radiation dose in small animals undergoing irradiation in an orthovoltage x-ray unit. Extensive research is being done with small animals, particularly mice and rats, in fields such as cancer therapy, radiation biology and radiological countermeasures. Results and conclusion are often drawn from research based solely on manufacturer's specifications of the delivered dose rate without independent verification or adequate understanding of the machines' capabilities. Accurate radiation dose information is paramount when conducting research in this arena. Traditional methods of dosimetry, namely thermoluminescence dosimeters (TLDs) are challenging and often time consuming. This section hopes to show that in place of TLDs, MOSFETs can provide accurate, precise dose information comparable with TLDs and ionization chambers. Measurements of all three dosimeters are compared in a small animal irradiator in phantoms and in vivo. Measurements done with MOSFETs are shown to deviate by 2.5% from that of the ADCL calibrated ionization chamber while TLDs showed a 7% deviation. Dose distributions within a phantom is also measured using radiochromic film to estimate the attenuation and show that dose is not uniform throughout the mouse. A dose decrease of approximately 30% is observed in a water phantom, which was only slightly mitigated by a hardening the beam with additional filtration. A Bland-Altman plot was created to show that the MOSFETs and TLDs used to make the dose measurements are statistically equivalent. The results show that all measurements made over a range of doses fall within 1.96 standard deviations of the mean.

  14. Placement of iliosacral screws using 3D image-guided (O-Arm) technology and Stealth Navigation: comparison with traditional fluoroscopy.

    PubMed

    Theologis, A A; Burch, S; Pekmezci, M

    2016-05-01

    We compared the accuracy, operating time and radiation exposure of the introduction of iliosacral screws using O-arm/Stealth Navigation and standard fluoroscopy. Iliosacral screws were introduced percutaneously into the first sacral body (S1) of ten human cadavers, four men and six women. The mean age was 77 years (58 to 85). Screws were introduced using a standard technique into the left side of S1 using C-Arm fluoroscopy and then into the right side using O-Arm/Stealth Navigation. The radiation was measured on the surgeon by dosimeters placed under a lead thyroid shield and apron, on a finger, a hat and on the cadavers. There were no neuroforaminal breaches in either group. The set-up time for the O-Arm was significantly longer than for the C-Arm, while total time for placement of the screws was significantly shorter for the O-Arm than for the C-Arm (p = 0.001). The mean absorbed radiation dose during fluoroscopy was 1063 mRad (432.5 mRad to 4150 mRad). No radiation was detected on the surgeon during fluoroscopy, or when he left the room during the use of the O-Arm. The mean radiation detected on the cadavers was significantly higher in the O-Arm group (2710 mRem standard deviation (sd) 1922) than during fluoroscopy (11.9 mRem sd 14.8) (p < 0.01). O-Arm/Stealth Navigation allows for faster percutaneous placement of iliosacral screws in a radiation-free environment for surgeons, albeit with the same accuracy and significantly more radiation exposure to cadavers, when compared with standard fluoroscopy. Placement of iliosacral screws with O-Arm/Stealth Navigation can be performed safely and effectively. Cite this article: Bone Joint J 2016;98-B:696-702. ©2016 The British Editorial Society of Bone & Joint Surgery.

  15. Endoscopic bronchial occlusion for postoperative persistent bronchopleural fistula with computed tomography fluoroscopy guidance and virtual bronchoscopic navigation

    PubMed Central

    Yanagiya, Masahiro; Matsumoto, Jun; Nagano, Masaaki; Kusakabe, Masashi; Matsumoto, Yoko; Furukawa, Ryutaro; Ohara, Sayaka; Usui, Kazuhiro

    2018-01-01

    Abstract Rationale: The development of postoperative bronchopleural fistula (BPF) remains a challenge in thoracic surgery. We herein report a case of BPF successfully treated with endoscopic bronchial occlusion under computed tomography (CT) fluoroscopy and virtual bronchoscopic navigation (VBN). Patient concerns: A 63-year-old man underwent right upper lobectomy with concomitant S6a subsegmentectomy for lung adenocarcinoma. On postoperative day 24, he complained of shaking chills with high fever. Diagnoses: BPF with subsequent pneumonia and empyema. Interventions: Despite aggressive surgical interventions for the BPF, air leakage persisted postoperatively. On days 26 and 34 after the final operation, endobronchial occlusions were performed under CT fluoroscopy and VBN. Outcomes: The air leaks greatly decreased and the patient was discharged. Lessons: CT fluoroscopy and VBN can be useful techniques for endobronchial occlusion in the treatment of BPF. PMID:29443771

  16. Initial clinical experience with a remote magnetic catheter navigation system for ablation of cavotricuspid isthmus-dependent right atrial flutter.

    PubMed

    Arya, Arash; Kottkamp, Hans; Piorkowski, Christopher; Bollmann, Andreas; Gerdes-Li, Jin-Hong; Riahi, Sam; Esato, Masahiro; Hindricks, Gerhard

    2008-05-01

    A remote magnetic navigation system (MNS) is available and has been used with a 4-mm-tip magnetic catheter for radiofrequency (RF) ablation of some supraventricular and ventricular arrhythmias; however, it has not been evaluated for the ablation of cavotricuspid isthmus-dependent right atrial flutter (AFL). The present study evaluates the feasibility and efficiency of this system and the newly available 8-mm-tip magnetic catheter to perform RF ablation in patients with AFL. Twenty-six consecutive patients (23 men, mean age 64.6 +/- 9.6 years) underwent RF ablation using a remote MNS. RF ablation was performed with an 8-mm-tip magnetic catheter (70 degrees C, maximum power 70 W, 90 seconds). The endpoint of ablation was complete bidirectional isthmus block. To assess a possible learning curve, procedural data were compared between the first 14 (group 1) and the rest (group 2) of the patients. The initial rhythm during ablation was AFL in 20 (19 counterclockwise and 1 clockwise) and sinus rhythm in six patients. Due to technical issues, the ablation in the 18th patient could not be done with the MNS, and so we switched to conventional ablation. The remote magnetic navigation and ablation procedure was successful in 24 of the 25 (96%) remaining patients with AFL. In one patient (patient 2), conventional catheter was used to complete the isthmus block after termination of AFL. The procedure, preparation, ablation, and fluoroscopy times (median [range]) were 53 (30-130) minutes, 28 (10-65) minutes, 25 (12-78) minutes, and 7.5 (3.2-20.8) minutes, respectively. Patients in group 2 had shorter procedure (45 [30-70] min vs 80 [57-130] min, P = 0.0001), preparation (25 [10-30] min vs 42 [30-65] min, P = 0.0001), ablation (20 [12-40] min vs 31 [20-78] min, P = 0.002), and fluoroscopy (7.2 [3.2-12.2] min vs 11.0 [5.4-20.8] min, P = 0.014) times. No complication occurred during the procedure. Using a remote MNS and an 8-mm-tip magnetic catheter, ablation of AFL is feasible, safe, and effective. Our data suggest that there is a short learning curve for this procedure.

  17. Radiation exposure to eye lens and operator hands during endovascular procedures in hybrid operating rooms.

    PubMed

    Attigah, Nicolas; Oikonomou, Kyriakos; Hinz, Ulf; Knoch, Thomas; Demirel, Serdar; Verhoeven, Eric; Böckler, Dittmar

    2016-01-01

    The purpose of this study was to evaluate the radiation exposure of vascular surgeons' eye lens and fingers during complex endovascular procedures in modern hybrid operating rooms. Prospective, nonrandomized multicenter study design. One hundred seventy-one consecutive patients (138 male; median age, 72.5 years [interquartile range, 65-77 years]) underwent an endovascular procedure in a hybrid operating room between March 2012 and July 2013 in two vascular centers. The dose-area product (DAP), fluoroscopy time, operating time, and amount of contrast dye were registered prospectively. For radiation dose recordings, single-use dosimeters were attached at eye level and to the ring finger of the hand next to the radiation field of the operator for each endovascular procedure. Dose recordings were evaluated by an independent institution. Before the study, precursory investigations were obtained to simulate the radiation dose to eye lens and fingers with an Alderson phantome (RSD, Long Beach, Calif). Interventions were classified into six treatment categories: endovascular repair of infrarenal abdominal aneurysm (n = 65), thoracic endovascular aortic repair (n = 32), branched endovascular aortic repair for thoracoabdominal aneurysms (n = 17), fenestrated endovascular aortic repair for complex abdominal aortic aneurysm, (n = 25), iliac branched device (n = 8), and peripheral interventions (n = 24). There was a significant correlation in DAP between both lens (P < .01; r = 0.55) and finger (P < .01; r = 0.56) doses. The estimated fluoroscopy time to reach a radiation threshold of 20 mSv/y was 1404.10 minutes (90% confidence limit, 1160, 1650 minutes). According to correlation of the lens dose with the DAP an estimated cumulative DAP of 932,000 mGy/m(2) (90% confidence limit, 822,000, 1,039,000) would be critical for a threshold of 20 mSv/y for the eyes. Radiation protection is a serious issue for vascular surgeons because most complex endovascular procedures are delivering measurable radiation to the eyes. With the correlation of the DAP obtained in standard endovascular procedures a critical threshold of 20 mSv/y to the eyes can be predicted and thus an estimate of a potential harmful exposure to the eyes can be obtained. Copyright © 2016 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

  18. Radiation safety in the cardiac catheterization lab: A time series quality improvement initiative.

    PubMed

    Abuzeid, Wael; Abunassar, Joseph; Leis, Jerome A; Tang, Vicky; Wong, Brian; Ko, Dennis T; Wijeysundera, Harindra C

    Interventional cardiologists have one of the highest annual radiation exposures yet systems of care that promote radiation safety in cardiac catheterization labs are lacking. This study sought to reduce the frequency of radiation exposure, for PCI procedures, above 1.5Gy in labs utilizing a Phillips system at our local institution by 40%, over a 12-month period. We performed a time series study to assess the impact of different interventions on the frequency of radiation exposure above 1.5Gy. Process measures were percent of procedures where collimation and magnification were used and percent of completion of online educational modules. Balancing measures were the mean number of cases performed and mean fluoroscopy time. Information sessions, online modules, policies and posters were implemented followed by the introduction of a new lab with a novel software (AlluraClarity©) to reduce radiation dose. There was a significant reduction (91%, p<0.05) in the frequency of radiation exposure above 1.5Gy after utilizing a novel software (AlluraClarity©) in a new Phillips lab. Process measures of use of collimation (95.0% to 98.0%), use of magnification (20.0% to 14.0%) and completion of online modules (62%) helped track implementation. The mean number of cases performed and mean fluoroscopy time did not change significantly. While educational strategies had limited impact on reducing radiation exposure, implementing a novel software system provided the most effective means of reducing radiation exposure. Crown Copyright © 2017. Published by Elsevier Inc. All rights reserved.

  19. Rotational Angiography Based Three-Dimensional Left Atrial Reconstruction: A New Approach for Transseptal Puncture.

    PubMed

    Koektuerk, Buelent; Yorgun, Hikmet; Koektuerk, Oezlem; Turan, Cem H; Gorr, Eduard; Horlitz, Marc; Turan, Ramazan G

    2016-02-01

    Rotational angiography is a well-known method for the three-dimensional (3-D) reconstruction of left atrium and pulmonary veins during left-sided atrial arrhythmia ablation procedures. In our study, we aimed to review our experience in transseptal puncture (TSP) using 3-D rotational angiography. We included a total of 271 patients who underwent atrial fibrillation ablation using cryoballoon. Rotational angiography was performed to get the three-dimensional left atrial and pulmonary vein reconstructions using cardiac C-arm computed tomography. The image reconstruction was made using the DynaCT Cardiac software (Siemens, Erlangen, Germany). The mean age of the study population was 61 ± 10 years. The indications for left atrial arrhythmia ablation were paroxysmal AF in 140 patients (52%) and persistent AF patients in 131 (48%) patients. The success rate of TSP using only rotational guidance was (264/271 patients, 97.4%). In the remaining seven patients, transesophageal guidance was used after the initial attempt due to thick interatrial septum in five patients and difficult TSP due to abnormal anatomy and mild pericardial effusion in the remaining two patients. Mean fluoroscopy dosage of the rotational angiography was 4896.4 ± 825.3 μGym(2). The mean time beginning from femoral vein puncture to TSP was 12.3 ± 5.5 min. TSP guided by rotational angiography is a safe and effective method. Our results indicate that integration of rotational angiographic images into the real-time fluoroscopy can guide the TSP during the procedure. © 2015 John Wiley & Sons Ltd.

  20. Ultrasound guidance to perform intra-articular injection of gadolinium-based contrast material for magnetic resonance arthrography as an alternative to fluoroscopy: the time is now.

    PubMed

    Messina, Carmelo; Banfi, Giuseppe; Aliprandi, Alberto; Mauri, Giovanni; Secchi, Francesco; Sardanelli, Francesco; Sconfienza, Luca Maria

    2016-05-01

    Magnetic resonance (MR) imaging has been definitively established as the reference standard in the evaluation of joints in the body. Similarly, magnetic resonance arthrography has emerged as a technique that has been proven to increase significantly the diagnostic performance if compared with conventional MR imaging, especially when dealing with fibrocartilage and articular cartilage abnormalities. Diluted gadolinium can be injected in the joint space using different approaches: under palpation using anatomic landmarks or using an imaging guidance, such as fluoroscopy, computed tomography, or ultrasound. Fluoroscopy has been traditionally used, but the involvement of ionizing radiation should represent a remarkable limitation of this modality. Conversely, ultrasound has emerged as a feasible, cheap, quick, and radiation-free modality that can be used to inject joints, with comparable accuracy of fluoroscopy. In the present paper, we discuss the advantages and disadvantages of using fluoroscopy or ultrasound in injecting gadolinium-based contrast agents in joints to perform magnetic resonance arthrography, also in view of the new EuroSAFE Imaging initiative promoted by the European Society of Radiology and the recent updates to the European Atomic Energy Community 2013/59 directive on the medical use of ionizing radiation. • Intra-articular contrast agent injection can be performed using different imaging modalities • Fluoroscopy is widely used, but uses ionizing radiation • Ultrasound is an accurate, quick, and radiation-free modality for joint injection • X-rays should be avoided when other radiation-free modalities can be used.

  1. Radiation monitoring in interventional cardiology: a requirement

    NASA Astrophysics Data System (ADS)

    Rivera, T.; Uruchurtu, E. S.

    2017-01-01

    The increasing of procedures using fluoroscopy in interventional cardiology procedures may increase medical and patients to levels of radiation that manifest in unintended outcomes. Such outcomes may include skin injury and cancer. The cardiologists and other staff members in interventional cardiology are usually working close to the area under examination and they receive the dose primarily from scattered radiation from the patient. Mexico does not have a formal policy for monitoring and recording the radiation dose delivered in hemodynamic establishments. Deterministic risk management can be improved by monitoring the radiation delivered from X-ray devices. The objective of this paper is to provide cardiologist, techniques, nurses, and all medical staff an information on DR levels, about X-ray risks and a simple a reliable method to control cumulative dose.

  2. Endoscopic ultrasound-guided biliary drainage

    PubMed Central

    Chavalitdhamrong, Disaya; Draganov, Peter V

    2012-01-01

    Endoscopic ultrasound (EUS)-guided biliary drainage has emerged as a minimally invasive alternative to percutaneous and surgical interventions for patients with biliary obstruction who had failed endoscopic retrograde cholangiopancreatography (ERCP). EUS-guided biliary drainage has become feasible due to the development of large channel curvilinear therapeutic echo-endoscopes and the use of real-time ultrasound and fluoroscopy imaging in addition to standard ERCP devices and techniques. EUS-guided biliary drainage is an attractive option because of its minimally invasive, single step procedure which provides internal biliary decompression. Multiple investigators have reported high success and low complication rates. Unfortunately, high quality prospective data are still lacking. We provide detailed review of the use of EUS for biliary drainage from the perspective of practicing endoscopists with specific focus on the technical aspects of the procedure. PMID:22363114

  3. Multicenter Off-Label Use of Nit-Occlud Coil in Retrograde Closure of Small Patent Ductus Arteriosus.

    PubMed

    Zanjani, Keyhan Sayadpour; Sobhy, Rodina; El-Kaffas, Rania; El-Sisi, Amal

    2017-04-01

    We studied the safety and efficacy of closing patent ductus arteriosus by Nit-Occlud coils via retrograde approach. This is a retrospective study of 46 attempts to close ducts by this method in two hospitals in Egypt and Iran. Ductus arteriosus was crossed by left or right Judkins or endhole catheters. The coil was delivered via the same catheter or the provided endhole catheter after exchange. The procedure was successful in 42 out of 46 attempts. Fluoroscopy and procedural times were significantly shorter when the catheter was not exchanged. This method is effective and safe for the closure of small ducts. Crossing the duct and delivering the coil by a left Judkins catheter is the easiest and fastest way to perform this method.

  4. Comparative Prospective Study Reporting Intraoperative Parameters, Pedicle Screw Perforation, and Radiation Exposure in Navigation-Guided versus Non-navigated Fluoroscopy-Assisted Minimal Invasive Transforaminal Lumbar Interbody Fusion

    PubMed Central

    Kundnani, Vishal; Dutta, Shumayou; Patel, Ankit; Mehta, Gaurav; Singh, Mahendra

    2018-01-01

    Study Design Prospective cohort study. Purpose To compare intraoperative parameters, radiation exposure, and pedicle screw perforation rate in navigation-guided versus non-navigated fluoroscopy-assisted minimal invasive transforaminal lumbar interbody fusion (MIS TLIF). Overview of Literature The poor reliability of fluoroscopy-guided instrumentation and growing concerns about radiation exposure have led to the development of navigation-guided instrumentation techniques in MIS TLIF. The literature evaluating the efficacy of navigation-guided MIS TLIF is scant. Methods Eighty-seven patients underwent navigation- or fluoroscopy-guided MIS TLIF for symptomatic lumbar/lumbosacral spondylolisthesis. Demographics, intraoperative parameters (surgical time, blood loss), and radiation exposure (sec/mGy/Gy.cm2 noted from C-arm for comparison only) were recorded. Computed tomography was performed in patients in the navigation and non-navigation groups at postoperative 12 months and reviewed by an independent observer to assess the accuracy of screw placement, perforation incidence, location, grade (Mirza), and critical versus non-critical neurological implications. Results Twenty-seven patients (male/female, 11/16; L4–L5/L5–S1, 9/18) were operated with navigation-guided MIS TLIF, whereas 60 (male/female, 25/35; L4–L5/L5–S1, 26/34) with conventional fluoroscopy-guided MIS TILF. The use of navigation resulted in reduced fluoroscopy usage (dose area product, 0.47 Gy.cm2 versus 2.93 Gy.cm2), radiation exposure (1.68 mGy versus 10.97 mGy), and fluoroscopy time (46.5 seconds versus 119.08 seconds), with p-values of <0.001. Furthermore, 96.29% (104/108) of pedicle screws in the navigation group were accurately placed (grade 0) (4 breaches, all grade I) compared with 91.67% (220/240) in the non-navigation group (20 breaches, 16 grade I+4 grade II; p=0.114). None of the breaches resulted in a corresponding neurological deficit or required revision. Conclusions Navigation guidance in MIS TLIF reduced radiation exposure, but the perforation status was not statistically different than that for the fluoroscopy-based technique. Thus, navigation in nondeformity cases is useful for significantly reducing the radiation exposure, but its ability to reduce pedicle screw perforation in nondeformity cases remains to be proven. PMID:29713413

  5. Endovascular aneurysm repair simulation can lead to decreased fluoroscopy time and accurately delineate the proximal seal zone.

    PubMed

    Kim, Ann H; Kendrick, Daniel E; Moorehead, Pamela A; Nagavalli, Anil; Miller, Claire P; Liu, Nathaniel T; Wang, John C; Kashyap, Vikram S

    2016-07-01

    The use of simulators for endovascular aneurysm repair (EVAR) is not widespread. We examined whether simulation could improve procedural variables, including operative time and optimizing proximal seal. For the latter, we compared suprarenal vs infrarenal fixation endografts, right femoral vs left femoral main body access, and increasing angulation of the proximal aortic neck. Computed tomography angiography was obtained from 18 patients who underwent EVAR at a single institution. Patient cases were uploaded to the ANGIO Mentor endovascular simulator (Simbionix, Cleveland, Ohio) allowing for three-dimensional reconstruction and adapted for simulation with suprarenal fixation (Endurant II; Medtronic Inc, Minneapolis, Minn) and infrarenal fixation (C3; W. L. Gore & Associates Inc, Newark, Del) deployment systems. Three EVAR novices and three experienced surgeons performed 18 cases from each side with each device in randomized order (n = 72 simulations/participant). The cases were stratified into three groups according to the degree of infrarenal angulation: 0° to 20°, 21° to 40°, and 41° to 66°. Statistical analysis used paired t-test and one-way analysis of variance. Mean fluoroscopy time for participants decreased by 48.6% (P < .0001), and total procedure time decreased by 33.8% (P < .0001) when initial cases were compared with final cases. When stent deployment accuracy was evaluated across all cases, seal zone coverage in highly angulated aortic necks was significantly decreased. The infrarenal device resulted in mean aortic neck zone coverage of 91.9%, 89.4%, and 75.4% (P < .0001 by one-way analysis of variance), whereas the suprarenal device yielded 92.9%, 88.7%, and 71.5% (P < .0001) for the 0° to 20°, 21° to 40°, and 41° to 66° cases, respectively. Suprarenal fixation did not increase seal zone coverage. The side of femoral access for the main body did not influence proximal seal zone coverage regardless of infrarenal angulation. Simulation of EVAR leads to decreased fluoroscopy times for novice and experienced operators. Side of femoral access did not affect precision of proximal endograft landing. The angulated aortic neck leads to decreased proximal seal zone coverage regardless of infrarenal or suprarenal fixation devices. Copyright © 2016 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

  6. Electromagnetic Real Time Navigation in the Region of the Posterior Pelvic Ring: An Experimental In-Vitro Feasibility Study and Comparison of Image Guided Techniques.

    PubMed

    Pishnamaz, Miguel; Wilkmann, Christoph; Na, Hong-Sik; Pfeffer, Jochen; Hänisch, Christoph; Janssen, Max; Bruners, Philipp; Kobbe, Philipp; Hildebrand, Frank; Schmitz-Rode, Thomas; Pape, Hans-Christoph

    2016-01-01

    Electromagnetic tracking is a relatively new technique that allows real time navigation in the absence of radiation. The aim of this study was to prove the feasibility of this technique for the treatment of posterior pelvic ring fractures and to compare the results with established image guided procedures. Tests were performed in pelvic specimens (Sawbones®) with standardized sacral fractures (Type Denis I or II). A gel matrix simulated the operative approach and a cover was used to disable visual control. The electromagnetic setup was performed by using a custom made carbon reference plate and a prototype stainless steel K-wire with an integrated sensor coil. Four different test series were performed: Group OCT: Optical navigation using preoperative CT-scans; group O3D: Optical navigation using intraoperative 3-D-fluoroscopy; group Fluoro: Conventional 2-D-fluoroscopy; group EMT: Electromagnetic navigation combined with a preoperative Dyna-CT. Accuracy of screw placement was analyzed by standardized postoperative CT-scan for each specimen. Operation time and intraoperative radiation exposure for the surgeon was documented. All data was analyzed using SPSS (Version 20, 76 Chicago, IL, USA). Statistical significance was defined as p< 0.05. 160 iliosacral screws were placed (40 per group). EMT resulted in a significantly higher incidence of optimal screw placement (EMT: 36/40) compared to the groups Fluoro (30/40; p< 0.05) and OCT (31/40; p< 0.05). Results between EMT and O3D were comparable (O3D: 37/40; n.s.). Also, the operation time was comparable between groups EMT and O3D (EMT 7.62 min vs. O3D 7.98 min; n.s.), while the surgical time was significantly shorter compared to the Fluoro group (10.69 min; p< 0.001) and the OCT group (13.3 min; p< 0.001). Electromagnetic guided iliosacral screw placement is a feasible procedure. In our experimental setup, this method was associated with improved accuracy of screw placement and shorter operation time when compared with the conventional fluoroscopy guided technique and compared to the optical navigation using preoperative CT-scans. Further studies are necessary to rule out drawbacks of this technique regarding ferromagnetic objects.

  7. Estimating the probability of rare events: addressing zero failure data.

    PubMed

    Quigley, John; Revie, Matthew

    2011-07-01

    Traditional statistical procedures for estimating the probability of an event result in an estimate of zero when no events are realized. Alternative inferential procedures have been proposed for the situation where zero events have been realized but often these are ad hoc, relying on selecting methods dependent on the data that have been realized. Such data-dependent inference decisions violate fundamental statistical principles, resulting in estimation procedures whose benefits are difficult to assess. In this article, we propose estimating the probability of an event occurring through minimax inference on the probability that future samples of equal size realize no more events than that in the data on which the inference is based. Although motivated by inference on rare events, the method is not restricted to zero event data and closely approximates the maximum likelihood estimate (MLE) for nonzero data. The use of the minimax procedure provides a risk adverse inferential procedure where there are no events realized. A comparison is made with the MLE and regions of the underlying probability are identified where this approach is superior. Moreover, a comparison is made with three standard approaches to supporting inference where no event data are realized, which we argue are unduly pessimistic. We show that for situations of zero events the estimator can be simply approximated with 1/2.5n, where n is the number of trials. © 2011 Society for Risk Analysis.

  8. Novel real-time tumor-contouring method using deep learning to prevent mistracking in X-ray fluoroscopy.

    PubMed

    Terunuma, Toshiyuki; Tokui, Aoi; Sakae, Takeji

    2018-03-01

    Robustness to obstacles is the most important factor necessary to achieve accurate tumor tracking without fiducial markers. Some high-density structures, such as bone, are enhanced on X-ray fluoroscopic images, which cause tumor mistracking. Tumor tracking should be performed by controlling "importance recognition": the understanding that soft-tissue is an important tracking feature and bone structure is unimportant. We propose a new real-time tumor-contouring method that uses deep learning with importance recognition control. The novelty of the proposed method is the combination of the devised random overlay method and supervised deep learning to induce the recognition of structures in tumor contouring as important or unimportant. This method can be used for tumor contouring because it uses deep learning to perform image segmentation. Our results from a simulated fluoroscopy model showed accurate tracking of a low-visibility tumor with an error of approximately 1 mm, even if enhanced bone structure acted as an obstacle. A high similarity of approximately 0.95 on the Jaccard index was observed between the segmented and ground truth tumor regions. A short processing time of 25 ms was achieved. The results of this simulated fluoroscopy model support the feasibility of robust real-time tumor contouring with fluoroscopy. Further studies using clinical fluoroscopy are highly anticipated.

  9. AAPM/RSNA physics tutorial for residents: physics of flat-panel fluoroscopy systems: Survey of modern fluoroscopy imaging: flat-panel detectors versus image intensifiers and more.

    PubMed

    Nickoloff, Edward Lee

    2011-01-01

    This article reviews the design and operation of both flat-panel detector (FPD) and image intensifier fluoroscopy systems. The different components of each imaging chain and their functions are explained and compared. FPD systems have multiple advantages such as a smaller size, extended dynamic range, no spatial distortion, and greater stability. However, FPD systems typically have the same spatial resolution for all fields of view (FOVs) and are prone to ghosting. Image intensifier systems have better spatial resolution with the use of smaller FOVs (magnification modes) and tend to be less expensive. However, the spatial resolution of image intensifier systems is limited by the television system to which they are coupled. Moreover, image intensifier systems are degraded by glare, vignetting, spatial distortions, and defocusing effects. FPD systems do not have these problems. Some recent innovations to fluoroscopy systems include automated filtration, pulsed fluoroscopy, automatic positioning, dose-area product meters, and improved automatic dose rate control programs. Operator-selectable features may affect both the patient radiation dose and image quality; these selectable features include dose level setting, the FOV employed, fluoroscopic pulse rates, geometric factors, display software settings, and methods to reduce the imaging time. © RSNA, 2011.

  10. Fluoroscopy and imageless navigation enable an equivalent reconstruction of leg length and global and femoral offset in THA.

    PubMed

    Weber, Markus; Woerner, Michael; Springorum, Robert; Sendtner, Ernst; Hapfelmeier, Alexander; Grifka, Joachim; Renkawitz, Tobias

    2014-10-01

    Restoration of biomechanics is a major goal in THA. Imageless navigation enables intraoperative control of leg length equalization and offset reconstruction. However, the effect of navigation compared with intraoperative fluoroscopy is unclear. We asked whether intraoperative use of imageless navigation (1) improves the relative accuracy of leg length and global and femoral offset restoration; (2) increases the absolute precision of leg length and global and femoral offset equalization; and (3) reduces outliers in a reconstruction zone of ± 5 mm for leg length and global and femoral offset restoration compared with intraoperative fluoroscopy during minimally invasive (MIS) THA with the patient in a lateral decubitus position. In this prospective study a consecutive series of 125 patients were randomized to either navigation-guided or fluoroscopy-controlled THA using sealed, opaque envelopes. All patients received the same cementless prosthetic components through an anterolateral MIS approach while they were in a lateral decubitus position. Leg length, global or total offset (representing the combination of femoral and acetabular offset), and femoral offset differences were restored using either navigation or fluoroscopy. Postoperatively, residual leg length and global and femoral offset discrepancies were analyzed on magnification-corrected radiographs of the pelvis by an independent and blinded examiner using digital planning software. Accuracy was defined as the relative postoperative difference between the surgically treated and the unaffected contralateral side for leg length and offset, respectively; precision was defined as the absolute postoperative deviation of leg length and global and femoral offset regardless of lengthening or shortening of leg length and offset throughout the THA. All analyses were performed per intention-to-treat. Analyzing the relative accuracy of leg length restoration we found a mean difference of 0.2 mm (95% CI, -1.0 to +1.4 mm; p = 0.729) between fluoroscopy and navigation, 0.2 mm (95 % CI, -0.9 to +1.3 mm; p = 0.740) for global offset and 1.7 mm (95 % CI, +0.4 to +2.9 mm; p = 0.008) for femoral offset. For the absolute precision of leg length and global and femoral offset equalization, there was a mean difference of 1.7 ± 0.3 mm (p < 0.001) between fluoroscopy and navigation. The biomechanical reconstruction with a residual leg length and global and femoral offset discrepancy less than 5 mm and less than 8 mm, respectively, succeeded in 93% and 98%, respectively, in the navigation group and in 54% and 95%, respectively, in the fluoroscopy group. Intraoperative fluoroscopy and imageless navigation seem equivalent in accuracy and precision to reconstruct leg length and global and femoral offset during MIS THA with the patient in the lateral decubitus position.

  11. Palliation double stenting for malignant biliary and duodenal obstruction

    PubMed Central

    ZHAO, LIANG; XU, HAITAO; ZHANG, YUBAO

    2016-01-01

    The surgical management of patients with malignant biliary and duodenal obstruction is complex. Tumor excision is no longer possible in the majority of patients with malignant obstructive jaundice and duodenal obstruction. The aim of the present study was to evaluate the effectiveness of intraluminal dual stent placement in malignant biliary and duodenal obstruction. In total, 20 patients with malignant obstructive jaundice and duodenal obstruction, including 6 with pancreatic carcinoma, 11 with cholangiocarcinoma, 1 with duodenal carcinoma and 2 with abdominal lymph node metastasis, were treated with intraluminal stent placement. Bile duct obstruction with late occurrence of duodenal obstruction was observed in 16 cases, and duodenal obstruction followed by a late occurrence of bile duct obstruction was observed in 3 cases, while, in 1 case, bile duct obstruction and duodenal obstruction occurred simultaneously. After X-ray fluoroscopy revealed obstruction in the bile duct and duodenum, stents were placed into the respective lumens. Percutaneous transhepatic placement was employed for the biliary stent, while the duodenal stent was placed perioraly. The clinical outcomes, including complications associated with the procedures and patency of the stents, were evaluated. The biliary and duodenal stents were successfully implanted in 18 patients and the technical success rate was 90% (18/20). A total of 39 stents were implanted in 20 patients. In 2 cases, duodenal stent placement failed following biliary stent placement. Duodenal obstruction remitted in 15 patients, and 1 patient succumbed to aspiration pneumonia 5 days after the procedure. No severe complications were observed in any other patient. The survival time of the 18 patients was 5–21 months (median, 9.6 months), and 6 of those patients survived for >12 months. The present study suggests that X-ray fluoroscopy-guided intraluminal stent implantation is an effective procedure for the treatment of malignant biliary and duodenal obstruction. PMID:26889267

  12. Occupational radiation exposure and its health effects on interventional medical workers: study protocol for a prospective cohort study.

    PubMed

    Ko, Seulki; Chung, Hwan Hoon; Cho, Sung Bum; Jin, Young Woo; Kim, Kwang Pyo; Ha, Mina; Bang, Ye Jin; Ha, Yae Won; Lee, Won Jin

    2017-12-15

    Although fluoroscopically guided procedures involve a considerably high dose of radiation, few studies have investigated the effects of radiation on medical workers involved in interventional fluoroscopy procedures. Previous research remains in the early stages and has not reached a level comparable with other occupational studies thus far. Furthermore, the study of radiation workers provides an opportunity to estimate health risks at low doses and dose rates of ionising radiation. Therefore, the objectives of this study are (1) to initiate a prospective cohort study by conducting a baseline survey among medical radiation workers who involve interventional fluoroscopy procedures and (2) to assess the effect of occupational radiation exposure and on the overall health status through an in-depth cross-sectional study. Intervention medical workers in Korea will be enrolled by using a self-administered questionnaire survey, and the survey data will be linked with radiation dosimetry data, National Health Insurance claims data, cancer registry and mortality data. After merging these data, the radiation organ dose, lifetime attributable risk due to cancer and the risk per unit dose will be estimated. For the cross-sectional study, approximately 100 intervention radiology department workers will be investigated for blood tests, clinical examinations such as ultrasonography (thyroid and carotid artery scan) and lens opacity, the validation of badge dose and biodosimetry. This study was reviewed and approved by the institutional review board of Korea University (KU-IRB-12-12-A-1). All participants will provide written informed consent prior to enrolment. The findings of the study will be disseminated through peer-reviewed scientific journals, conference presentations, and a report will be submitted to the relevant public health authorities in the Korea Centers for Disease Control and Prevention to help with the development of appropriate research and management policies. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2017. All rights reserved. No commercial use is permitted unless otherwise expressly granted.

  13. Catheter ablation of ventricular tachycardias using remote magnetic navigation: a consecutive case-control study.

    PubMed

    Szili-Torok, Tamas; Schwagten, Bruno; Akca, Ferdi; Bauernfeind, Tamas; Abkenari, Lara Dabiri; Haitsma, David; Van Belle, Yves; Groot, Natasja D E; Jordaens, Luc

    2012-09-01

    Remote Magnetic Navigation for VT Ablation. This study aimed to compare acute and late outcomes of VT ablation using the magnetic navigation system (MNS) to manual techniques (MAN) in patients with (SHD) and without (NSHD) structural heart disease. Ablation data of 113 consecutive patients (43 SHD, 70 NSHD) with ventricular tachycardia treated with catheter ablation at our center were analyzed. Success rate, complications, procedure, fluoroscopy, and ablation times, and recurrence rates were systematically recorded for all patients. A total of 72 patients were included in the MNS group and 41 patients were included in the MAN group. Patient age, gender, and right ventricular and left ventricular VT were equally distributed. Acute success was achieved in 59 patients in the MNS group (82%) versus 27 (66%) patients in the MAN group (P = 0.046). Overall procedural time (177 ± 79 vs 232 ± 99 minutes, P < 0.01) and mean patient fluoroscopy time (27 ± 19 vs 56 ± 32 minutes, P < 0.001) were all significantly lower using MNS. In NSHD pts, higher acute success was achieved with MNS (83,7% vs 61.9%, P = 0.049), with shorter procedure times (151 ± 57 vs 210 ± 96, P = 0.011), whereas in SHD-VT these were not significantly different. No major complications occurred in the MNS group (0%) versus 1 cardiac tamponade and 1 significantly damaged ICD lead in the MAN group (4.9%, NS). After follow-up (20 ± 11 vs 20 ± 10 months, NS), VT recurred in 14 pts (23.7%) in the MNS group versus 12 pts (44.4%) in the MAN group (P = 0.047). The use of MNS offers advantages for ablation of NSHD-VT, while it offers similar efficacy for SHD-VT. (J Cardiovasc Electrophysiol, Vol. 23, pp. 948-954, September 2012). © 2012 Wiley Periodicals, Inc.

  14. Magnetically assisted remote-controlled endovascular catheter for interventional MR imaging: in vitro navigation at 1.5 T versus X-ray fluoroscopy.

    PubMed

    Losey, Aaron D; Lillaney, Prasheel; Martin, Alastair J; Cooke, Daniel L; Wilson, Mark W; Thorne, Bradford R H; Sincic, Ryan S; Arenson, Ronald L; Saeed, Maythem; Hetts, Steven W

    2014-06-01

    To compare in vitro navigation of a magnetically assisted remote-controlled (MARC) catheter under real-time magnetic resonance (MR) imaging with manual navigation under MR imaging and standard x-ray guidance in endovascular catheterization procedures in an abdominal aortic phantom. The 2-mm-diameter custom clinical-grade microcatheter prototype with a solenoid coil at the distal tip was deflected with a foot pedal actuator used to deliver 300 mA of positive or negative current. Investigators navigated the catheter into branch vessels in a custom cryogel abdominal aortic phantom. This was repeated under MR imaging guidance without magnetic assistance and under conventional x-ray fluoroscopy. MR experiments were performed at 1.5 T by using a balanced steady-state free precession sequence. The mean procedure times and percentage success data were determined and analyzed with a linear mixed-effects regression analysis. The catheter was clearly visible under real-time MR imaging. One hundred ninety-two (80%) of 240 turns were successfully completed with magnetically assisted guidance versus 144 (60%) of 240 turns with nonassisted guidance (P < .001) and 119 (74%) of 160 turns with standard x-ray guidance (P = .028). Overall mean procedure time was shorter with magnetically assisted than with nonassisted guidance under MR imaging (37 seconds ± 6 [standard error of the mean] vs 55 seconds ± 3, P < .001), and time was comparable between magnetically assisted and standard x-ray guidance (37 seconds ± 6 vs 44 seconds ± 3, P = .045). When stratified by angle of branch vessel, magnetic assistance was faster than nonassisted MR guidance at turns of 45°, 60°, and 75°. In this study, a MARC catheter for endovascular navigation under real-time MR imaging guidance was developed and tested. For catheterization of branch vessels arising at large angles, magnetically assisted catheterization was faster than manual catheterization under MR imaging guidance and was comparable to standard x-ray guidance.

  15. Catheter ablation of atrial fibrillation using remote magnetic catheter navigation: a case-control study.

    PubMed

    Arya, Arash; Zaker-Shahrak, Ruzbeh; Sommer, Phillip; Bollmann, Andreas; Wetzel, Ulrike; Gaspar, Thomas; Richter, Sergio; Husser, Daniela; Piorkowski, Christopher; Hindricks, Gerhard

    2011-01-01

    To compare the acute and the 6 month outcome of catheter ablation of atrial fibrillation (AF) using irrigated tip magnetic catheter and remote magnetic cathter navigation (RMN) with manual catheter navigation (MCN) in patients with paroxysmal and persistent AF. In this retrospective analysis 356 patients (235 male, mean age: 57.9 ± 10.9 years) with AF (70.5%, paroxysmal) who underwent catheter ablation between August 2007 and May 2008 using either RMN (n = 70, 46 male, mean age: 57.9 ± 10.1 years, 50% paroxysmal) or MCN (n = 286, 189 male, mean age: 58.0 ± 13.9 years, 75.5% paroxysmal) were included. All patients completed an intensive follow-up strategy. Complete pulmonary vein isolation was achieved in 87.6 and 99.6% of patients in RMN and MCN groups, respectively (P < 0.05). The procedure, fluoroscopy, and radiofrequency application times were 223 ± 44 vs. 166 ± 52 min (P < 0.0001), 13.7 ± 7.8 vs. 34.5 ± 15.1 min (P < 0.0001), and 75.4 ± 20.9 vs. 53.2 ± 21.4 min (P < 0.0001) in RMN and MCN groups, respectively. Seven (10.0%) and 28 (9.8%) patients in RMN and MCN groups received antiarrhythmic medications during the follow-up (P = 0.96). All the patients completed the 6 month follow-up. Freedom from AF at 6 months was achieved in 57.8 and 66.4% of the patients in RMN and MCN groups, respectively (P = 0.196). In patients without previous AF catheter ablation procedure the freedom from AF at 6 months were 68.2 and 60.5% in the MCN and RMN groups, respectively (P = 0.36). Catheter ablation using irrigated tip magnetic catheter and RMN is an effective and safe method for catheter ablation of AF. Compared to manual catheter navigation, the procedure and radiofrequency application times were longer and fluoroscopy time was shorter in the RMN group compared with the MCN group.

  16. Long-term efficacy of single procedure remote magnetic catheter navigation for ablation of ischemic ventricular tachycardia: a retrospective study.

    PubMed

    Dinov, Borislav; Schönbauer, Robert; Wojdyla-Hordynska, Agnieska; Braunschweig, Frieder; Richter, Sergio; Altmann, David; Sommer, Philipp; Gaspar, Thomas; Bollmann, Andreas; Wetzel, Ulrike; Rolf, Sascha; Piorkowski, Christopher; Hindricks, Gerhard; Arya, Arash

    2012-05-01

    Remote magnetic navigation (RMN) aims to reduce some inherent limitations of manual radiofrequency (RF) ablation. However, data comparing the effectiveness of both methods are scarce. This study evaluated the acute and long-term success of RMN guided versus manual RF ablation in patients with ischemic sustained ventricular tachycardia (sVT). One hundred two consecutive patients (age 68 ± 10 years, LVEF 32 ± 12%, 88 men) with ischemic sVT were ablated with RMN (Stereotaxis; 49%) or manually (51%) using substrate and/or activation mapping (Carto) and open-irrigated-tip catheters. All received implantable defibrillators or loop recorders. Acute success was defined as noninducibility of any sVT at the end of the ablation procedure and long-term success as freedom from VT upon follow-up. There was no difference in the baseline characteristics between the groups. Three patients died in hospital. Acute success rate was similar for RMN and manual ablation (82% vs 71%, P = 0.246). RMN was associated with significantly shorter fluoroscopy time (13 ± 12 minutes vs 32 ± 17 minutes, P = 0.0001) and RF time (2337.59 ± 1248.22 seconds vs 1589.95 ± 1047.42 seconds, P = 0.049), although total procedure time was similar (157 ± 40 minutes vs 148 ± 50 minutes, P = 0.42). There was a nonsignificant trend toward better long-term success in RMN group: after a median of 13 (range 1-34) months, 63% in the RMN and 53% in the manual ablation group were free from VT recurrence (P = 0.206). RMN guided RF ablation of ischemic sustained VT is equally efficient compared with manual ablation in terms of acute and long-term success rate. These results are achieved with a significantly reduced fluoroscopy time and shorter RF time. © 2012 Wiley Periodicals, Inc.

  17. Efficacy and safety of cryoballoon ablation versus radiofrequency catheter ablation in atrial fibrillation: an updated meta-analysis

    PubMed Central

    Ma, Honglan; Sun, Dongdong; Luan, Hui; Feng, Wei; Zhou, Yaqiong; Wu, Jine; He, Caiyun

    2017-01-01

    Introduction Cryoballoon ablation (CBA) and irrigated radiofrequency catheter ablation (RFCA) are the main treatments for drug-refractory symptomatic atrial fibrillation (AF). Aim To compare the efficacy and safety between CBA and RFCA for the treatment of AF. Material and methods We searched the Embase and Medline databases for clinical studies published up to December 2016. Studies that satisfied our predefined inclusion criteria were included. Results After searching through the literature in the two major databases, 20 studies with a total of 9,141 patients were included in our study. The CBA had a significantly shorter procedure time (weighted mean difference (WMD) –30.38 min; 95% CI: –46.43 to –14.33, p = 0.0002) and non-significantly shorter fluoroscopy time (WMD –3.18 min; 95% CI: –6.43 to 0.07, p = 0.06) compared with RFCA. There was no difference in freedom from AF between CBA and RFCA (CBA 78.55% vs. RFCA 83.13%, OR = 1.15, 95% CI: 0.95–1.39, p = 0.14). The CBA was associated with a high risk of procedure-related complications (CBA 9.02% vs. RFCA 6.56%, OR = 1.56, 95% CI: 1.05–2.31, p = 0.03), especially phrenic nerve paralysis (PNP, OR = 10.72, 95% CI: 5.59–20.55, p < 0.00001). The risk of pericardial effusions/cardiac tamponade was low in the CBA group (CBA 1.05% vs. RFCA 1.86%, OR = 0.62, 95% CI: 0.41–0.93, p = 0.02). Conclusions For AF, CBA was as effective as RFCA. However, CBA had a shorter procedure time and a non-significantly shorter fluoroscopy time, a significantly high risk of PNP and a low incidence of pericardial effusions/cardiac tamponade compared with RFCA. PMID:29056997

  18. Time-to-isolation guided titration of freeze duration in 3rd generation short-tip cryoballoon pulmonary vein isolation - Comparable clinical outcome and shorter procedure duration.

    PubMed

    Pott, Alexander; Kraft, Christoph; Stephan, Tilman; Petscher, Kerstin; Rottbauer, Wolfgang; Dahme, Tillman

    2018-03-15

    The optimal freeze duration in cryoballoon pulmonary vein isolation (PVI) is unknown. The 3rd generation cryoballoon facilitates observation of the time-to-isolation (TTI) and thereby enables individualized cryoenergy titration. To evaluate the efficacy of an individualized freeze duration we compared the clinical outcome of patients treated with a TTI-guided ablation protocol to the outcome of patients treated with a fixed ablation protocol. We compared 100 patients treated with the 3rd generation cryoballoon applying a TTI-based protocol (TTI group) to 100 patients treated by a fixed freeze protocol (fixed group). In the fixed group a 240s freeze cycle was followed by a 240s bonus freeze after acute PV isolation. In the TTI group freeze duration was 180s if TTI was ≥30s and reduced to only 120s, if TTI was <30s. In case of a TTI >60s a 180s bonus freeze was applied. Freedom from atrial arrhythmia recurrence off class I/III antiarrhythmic drugs after one year was not different between the TTI group (73.6%) and the fixed group (75.7%; p=0.75). Mean procedure duration was 85.8±27.3min in the TTI group compared to 115.7±27.1min in the fixed group (p<0.001). Mean fluoroscopy time was 17.5±6.6min in the TTI group and 22.5±9.8min in the fixed group (p<0.001). TTI-guided cryoenergy titration leads to reduced procedure duration and fluoroscopy time and appears to be as effective as a fixed ablation strategy. A single 2-minute freeze seems to be sufficient in case of short TTI. Copyright © 2017 The Authors. Published by Elsevier B.V. All rights reserved.

  19. SU-F-P-44: A Direct Estimate of Peak Skin Dose for Interventional Fluoroscopy Procedures

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Weir, V; Zhang, J

    Purpose: There is an increasing demand for medical physicist to calculate peak skin dose (PSD) for interventional fluoroscopy procedures. The dose information (Dose-Area-Product and Air Kerma) displayed in the console cannot directly be used for this purpose. Our clinical experience shows that the use of the existing methods may overestimate or underestimate PSD. This study attempts to develop a direct estimate of PSD from the displayed dose metrics. Methods: An anthropomorphic torso phantom was used for dose measurements for a common fluoroscopic procedure. Entrance skin doses were measured with a Piranha solid state point detector placed on the table surfacemore » below the torso phantom. An initial “reference dose rate” (RE) measurement was conducted by comparing the displayed dose rate (mGy/min) to the dose rate measured. The distance from table top to focal spot was taken as the reference distance (RD at the RE. Table height was then adjusted. The displayed air kerma and DAP were recorded and sent to three physicists to estimate PSD. An inverse square correction was applied to correct displayed air kerma at various table heights. The PSD estimated by physicists and the PSD by the proposed method were then compared with the measurements. The estimated DAPs were compared to displayed DAP readings (mGycm2). Results: The difference between estimated PSD by the proposed method and direct measurements was less than 5%. For the same set of data, the estimated PSD by each of three physicists is different from measurements by ±52%. The DAP calculated by the proposed method and displayed DAP readings in the console is less than 20% at various table heights. Conclusion: PSD may be simply estimated from displayed air kerma or DAP if the distance between table top and tube focal spot or if x-ray beam area on table top is available.« less

  20. Robotic-Assisted Versus Manual Prostatic Arterial Embolization for Benign Prostatic Hyperplasia: A Comparative Analysis

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Bagla, Sandeep, E-mail: sandeep.bagla@gmail.com; Smirniotopoulos, John; Orlando, Julie C.

    PurposeProstatic artery embolization (PAE) is a safe and efficacious procedure for benign prostatic hyperplasia (BPH), though is technically challenging. We present our experience of technical and clinical outcomes of robotic and manual PAE in patients with BPH.Materials and MethodsIRB-approved retrospective study of 40 consecutive patients 49–81 years old with moderate or severe grade BPH from May 2014 to July 2015: 20 robotic-assisted PAE (group 1), 20 manual PAE (group 2). Robotic-assisted PAE was performed using the Magellan Robotic System. American Urological Association (AUA-SI) score, cost, technical and clinical success, radiation dose, fluoroscopy, and procedure time were reviewed. Statistical analysis was performedmore » within and between each group using paired t test and one-way analysis of variance respectively, at 1 and 3 months.ResultsNo significant baseline differences in age and AUA-SI between groups. Technical success was 100% (group 1) and 95% (group 2). One unsuccessful subject from group 2 returned for a successful embolization using robotic assistance. Fluoroscopy and procedural times were similar between groups, with a non-significant lower patient radiation dose in group 1 (30,632.8 mGy/cm{sup 2} vs 35,890.9, p = 0.269). Disposable cost was significantly different between groups with the robotic-assisted PAE incurring a higher cost (group 1 $4530.2; group 2 $1588.5, p < 0.0001). Clinical improvement was significant in both arms at 3 months: group 1 mean change in AUA-SI of 8.3 (p = 0.006), group 2: 9.6 (p < 0.0001). No minor or major complications occurred.ConclusionsRobotic-assisted PAE offers technical success comparable to manual PAE, with similar clinical improvement with an increased cost.« less

  1. Occupational radiation exposure and its health effects on interventional medical workers: study protocol for a prospective cohort study

    PubMed Central

    Ko, Seulki; Chung, Hwan Hoon; Cho, Sung Bum; Jin, Young Woo; Kim, Kwang Pyo; Ha, Mina; Bang, Ye Jin; Ha, Yae Won; Lee, Won Jin

    2017-01-01

    Introduction Although fluoroscopically guided procedures involve a considerably high dose of radiation, few studies have investigated the effects of radiation on medical workers involved in interventional fluoroscopy procedures. Previous research remains in the early stages and has not reached a level comparable with other occupational studies thus far. Furthermore, the study of radiation workers provides an opportunity to estimate health risks at low doses and dose rates of ionising radiation. Therefore, the objectives of this study are (1) to initiate a prospective cohort study by conducting a baseline survey among medical radiation workers who involve interventional fluoroscopy procedures and (2) to assess the effect of occupational radiation exposure and on the overall health status through an in-depth cross-sectional study. Methods and analysis Intervention medical workers in Korea will be enrolled by using a self-administered questionnaire survey, and the survey data will be linked with radiation dosimetry data, National Health Insurance claims data, cancer registry and mortality data. After merging these data, the radiation organ dose, lifetime attributable risk due to cancer and the risk per unit dose will be estimated. For the cross-sectional study, approximately 100 intervention radiology department workers will be investigated for blood tests, clinical examinations such as ultrasonography (thyroid and carotid artery scan) and lens opacity, the validation of badge dose and biodosimetry. Ethics and dissemination This study was reviewed and approved by the institutional review board of Korea University (KU-IRB-12-12-A-1). All participants will provide written informed consent prior to enrolment. The findings of the study will be disseminated through peer-reviewed scientific journals, conference presentations, and a report will be submitted to the relevant public health authorities in the Korea Centers for Disease Control and Prevention to help with the development of appropriate research and management policies. PMID:29248885

  2. Echocardiographic demonstration of intracardiac glue after endoscopic obturation of gastroesophageal varices.

    PubMed

    Gallet, B; Zemour, G; Saudemont, J P; Renard, P; Hillion, M L; Hiltgen, M

    1995-01-01

    Systemic embolism is an unusual complication of endoscopic obturation of gastroesophageal varices with glue. This report describes a case of cerebral embolism after this procedure. Intracardiac glue within the left atrium was demonstrated by echocardiography. Cardiac fluoroscopy demonstrated an abnormal vessel connecting periesophageal veins with the right upper pulmonary vein. Cardiac surgery was performed. Intracardiac glue was removed and the entering orifice of the abnormal vessel in the right upper pulmonary vein was sutured. To our knowledge, this is the first reported case of intracardiac glue after variceal obturation. Echocardiography is useful in the diagnosis of this rare complication.

  3. Initial clinical experience with cardiac resynchronization therapy utilizing a magnetic navigation system.

    PubMed

    Gallagher, Peter; Martin, Laura; Angel, Lori; Tomassoni, Gery

    2007-02-01

    The placement of left ventricular (LV) leads during cardiac resynchronization therapy (CRT) involves many technical difficulties. These difficulties increase procedural times and decrease procedural success rates. A total of 50 patients with severe cardiomyopathy (mean LV ejection fraction was 21 +/- 6%) and a wide QRS underwent CRT implantation. Magnetic navigation (Stereotaxis, Inc.) was used to position a magnet-tipped 0.014'' guidewire (Cronus guidewire) within the coronary sinus (CS) vasculature. LV leads were placed in a lateral CS branch, either using a standard CS delivery sheath or using a "bare-wire" approach without a CS delivery sheath. The mean total procedure time was 98.1 +/- 29.1 minutes with a mean fluoroscopy time of 22.7 +/- 15.1 minutes. The mean LV lead positioning time was 10.4 +/- 7.6 minutes. The use of a delivery sheath was associated with longer procedure times 98 +/- 32 minutes vs 80 +/- 18 minutes (P = 0.029), fluoroscopy times 23 +/- 15 minutes vs 13 +/- 4 minutes (P = 0.0007) and LV lead positioning times 10 +/- 6 minutes vs 4 +/- 2 minutes (P = 0.015) when compared to a "bare-wire" approach. When compared with 52 nonmagnetic-assisted control CRT cases, magnetic navigation reduced total LV lead positioning times (10.4 +/- 7.6 minutes vs 18.6 +/- 18.9 minutes; P = 0.005). If more than one CS branch vessel was tested, magnetic navigation was associated with significantly shorter times for LV lead placement (16.2 +/- 7.7 minutes vs 36.4 +/- 23.4 minutes; P = 0.004). Magnetic navigation is a safe, feasible, and efficient tool for lateral LV lead placement during CRT. Magnetic navigation during CRT allows for control of the tip direction of the Cronus 0.014'' guidewire using either a standard CS delivery sheath or "bare-wire" approach. Although there are some important limitations to the 0.014'' Cronus magnetic navigation can decrease LV lead placement times compared with nonmagnetic-assisted control CRT cases, particularly if multiple CS branches are to be tested.

  4. Troponin T elevation after permanent pacemaker implantation.

    PubMed

    Chen, Xueying; Yu, Ziqing; Bai, Jin; Hu, Shulan; Wang, Wei; Qin, Shengmei; Wang, Jingfeng; Sun, Zhe; Su, Yangang; Ge, Junbo

    2017-08-01

    The objective of the study is to study the incidence, significance, and factors associated with cardiac troponin T (CTNT) elevation after pacemaker implantation. Three hundred seventy-four patients (104 single-chamber pacemakers or ICD, 243 dual-chamber pacemakers, and 27 cardiac resynchronization therapy/cardiac resynchronization therapy defibrillator) who had normal levels of CTNT at baseline and underwent implantation of a permanent pacemaker system were included in this study. Serum levels of CTNT were measured at baseline, 6 and 24 h after the implantation procedure. The median of CTNT levels increased from 0.012 ng/mL at baseline to 0.032 and 0.019 ng/mL at 6 and 24 h after the procedure, respectively (all p < 0.0001). Elevated CTNT levels were noted in 208 patients (55.6%) at 6 h after the implantation, among which 29 patients (7.8%) had CTNT levels exceeding the range of minimal myocardial damage (>0.09 ng/mL). After 1-year follow-up, the incidence of complications including dislodgement of the lead, pocket infection, pneumothorax, hemothorax, and vein thrombus and cardiac outcomes including hospitalization of heart failure, coronary artery disease, arrhythmia, and cardiovascular mortality was not significantly different between the normal and elevated CTNT groups at 6 h after the procedure. By logistic regression analysis, gender, N-terminal pro-B type natriuretic peptide (NT-pro-BNP) at baseline, left ventricular ejection fractions (LVEF), estimated glomerular filtration rate (eGFR), and fluoroscopy time were independently associated with CTNT elevation after adjusted for age, pacemaker types, right ventricle lead location (RVA or RVOT), heart function, and left ventricular end systolic dimension. Pacemaker implantation was found to be accompanied with CTNT elevation in 55.6% of the patients at 6 h after the procedure, and its kinetics were fast, which might not be related to the complications and adverse cardiac outcomes within 1 year of follow-up. Moreover, gender, NT-pro-BNP at baseline, LVEF, eGFR, and fluoroscopy time were found to be independent predictors of CTNT elevation.

  5. An integrated platform for image-guided cardiac resynchronization therapy

    NASA Astrophysics Data System (ADS)

    Ma, Ying Liang; Shetty, Anoop K.; Duckett, Simon; Etyngier, Patrick; Gijsbers, Geert; Bullens, Roland; Schaeffter, Tobias; Razavi, Reza; Rinaldi, Christopher A.; Rhode, Kawal S.

    2012-05-01

    Cardiac resynchronization therapy (CRT) is an effective procedure for patients with heart failure but 30% of patients do not respond. This may be due to sub-optimal placement of the left ventricular (LV) lead. It is hypothesized that the use of cardiac anatomy, myocardial scar distribution and dyssynchrony information, derived from cardiac magnetic resonance imaging (MRI), may improve outcome by guiding the physician for optimal LV lead positioning. Whole heart MR data can be processed to yield detailed anatomical models including the coronary veins. Cine MR data can be used to measure the motion of the LV to determine which regions are late-activating. Finally, delayed Gadolinium enhancement imaging can be used to detect regions of scarring. This paper presents a complete platform for the guidance of CRT using pre-procedural MR data combined with live x-ray fluoroscopy. The platform was used for 21 patients undergoing CRT in a standard catheterization laboratory. The patients underwent cardiac MRI prior to their procedure. For each patient, a MRI-derived cardiac model, showing the LV lead targets, was registered to x-ray fluoroscopy using multiple views of a catheter looped in the right atrium. Registration was maintained throughout the procedure by a combination of C-arm/x-ray table tracking and respiratory motion compensation. Validation of the registration between the three-dimensional (3D) roadmap and the 2D x-ray images was performed using balloon occlusion coronary venograms. A 2D registration error of 1.2 ± 0.7 mm was achieved. In addition, a novel navigation technique was developed, called Cardiac Unfold, where an entire cardiac chamber is unfolded from 3D to 2D along with all relevant anatomical and functional information and coupled to real-time device detection. This allowed more intuitive navigation as the entire 3D scene was displayed simultaneously on a 2D plot. The accuracy of the unfold navigation was assessed off-line using 13 patient data sets by computing the registration error of the LV pacing lead electrodes which was found to be 2.2 ± 0.9 mm. Furthermore, the use of Unfold Navigation was demonstrated in real-time for four clinical cases.

  6. Comparison of CT Fluoroscopy-Guided Manual and CT-Guided Robotic Positioning System for In Vivo Needle Placements in Swine Liver

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Cornelis, F.; Takaki, H.; Laskhmanan, M.

    PurposeTo compare CT fluoroscopy-guided manual and CT-guided robotic positioning system (RPS)-assisted needle placement by experienced IR physicians to targets in swine liver.Materials and MethodsManual and RPS-assisted needle placement was performed by six experienced IR physicians to four 5 mm fiducial seeds placed in swine liver (n = 6). Placement performance was assessed for placement accuracy, procedure time, number of confirmatory scans, needle manipulations, and procedure radiation dose. Intra-modality difference in performance for each physician was assessed using paired t test. Inter-physician performance variation for each modality was analyzed using Kruskal–Wallis test.ResultsPaired comparison of manual and RPS-assisted placements to a target by the samemore » physician indicated accuracy outcomes was not statistically different (manual: 4.53 mm; RPS: 4.66 mm; p = 0.41), but manual placement resulted in higher total radiation dose (manual: 1075.77 mGy/cm; RPS: 636.4 mGy/cm; p = 0.03), required more confirmation scans (manual: 6.6; RPS: 1.6; p < 0.0001) and needle manipulations (manual: 4.6; RPS: 0.4; p < 0.0001). Procedure time for RPS was longer than manual placement (manual: 6.12 min; RPS: 9.7 min; p = 0.0003). Comparison of inter-physician performance during manual placement indicated significant differences in the time taken to complete placements (p = 0.008) and number of repositions (p = 0.04) but not in other study measures (p > 0.05). Comparison of inter-physician performance during RPS-assisted placement suggested statistically significant differences in procedure time (p = 0.02) and not in other study measures (p > 0.05).ConclusionsCT-guided RPS-assisted needle placement reduced radiation dose, number of confirmatory scans, and needle manipulations when compared to manual needle placement by experienced IR physicians, with equivalent accuracy.« less

  7. The Advantage of a Ureteroscopic Navigation System with Magnetic Tracking in Comparison with Simulated Fluoroscopy in a Phantom Study.

    PubMed

    Yoshida, Kenji; Yokomizo, Akira; Matsuda, Tadashi; Hamasaki, Tsutomu; Kondo, Yukihiro; Yamaguchi, Kunihisa; Kanayama, Hiro-Omi; Wakumoto, Yoshiaki; Horie, Shigeo; Naito, Seiji

    2015-09-01

    To assess whether our ureteroscopic real-time navigation system has the possibility to reduce radiation exposure and improve performance of ureteroscopic maneuvers in surgeons of various ages and experience levels. Our novel ureteroscopic navigation system used a magnetic tracking device to detect the position of the ureteroscope and display it on a three-dimensional image. We recruited 31 urologists from five institutions to perform two tasks. Task 1 consisted of finding three internal markings on the phantom calices. Task 2 consisted of identifying all calices by ureteroscopy. In both tasks, participants performed with simulated fluoroscopy first, followed by our navigation system. Accuracy rates (AR) for identification, required time (T) for completing the task, migration length (ML), and time exposed to simulated fluoroscopy were recorded. The AR, T, and ML for both tasks were significantly better with the navigation system than without it (Task 1 with simulated fluoroscopy vs with navigation: AR 87.1 % vs 98.9%, P=0.003; T 355 s vs 191 s, P<0.0001; ML 4627 mm vs 2701 mm, P<0.0001. Task 2: AR 88.2% vs 96.7%, P=0.011; T 394 s vs 333 s, P=0.027; ML 5966 mm vs 5299 mm, P=0.0006). In both tasks, the participants used the simulated fluoroscopy about 20% of the total task time. Our navigation system, while still under development, could help surgeons of all levels to achieve better performances for ureteroscopic maneuvers compared with using fluoroscopic guidance. It also has the potential to reduce radiation exposure during fluoroscopy.

  8. CT Fluoroscopy-Guided Lung Biopsy with Novel Steerable Biopsy Canula: Ex-Vivo Evaluation in Ventilated Porcine Lung Explants

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Schaefer, Philipp J., E-mail: jp.schaefer@rad.uni-kiel.de; Fabel, Michael; Bolte, Hendrik

    2010-08-15

    The purpose was to evaluate ex-vivo a prototype of a novel biopsy canula under CT fluoroscopy-guidance in ventilated porcine lung explants in respiratory motion simulations. Using an established chest phantom for porcine lung explants, n = 24 artificial lesions consisting of a fat-wax-Lipiodol mixture (approx. 70HU) were placed adjacent to sensible structures such as aorta, pericardium, diaphragm, bronchus and pulmonary artery. A piston pump connected to a reservoir beneath a flexible silicone reconstruction of a diaphragm simulated respiratory motion by rhythmic inflation and deflation of 1.5 L water. As biopsy device an 18-gauge prototype biopsy canula with a lancet-like, helicallymore » bended cutting edge was used. The artificial lesions were punctured under CT fluoroscopy-guidance (SOMATOM Sensation 64, Siemens, Erlangen, Germany; 30mAs/120 kV/5 mm slice thickness) implementing a dedicated protocol for CT fluoroscopy-guided lung biopsy. The mean-diameter of the artificial lesions was 8.3 {+-} 2.6 mm, and the mean-distance of the phantom wall to the lesions was 54.1 {+-} 13.5 mm. The mean-displacement of the lesions by respiratory motion was 14.1 {+-} 4.0 mm. The mean-duration of CT fluoroscopy was 9.6 {+-} 5.1 s. On a 4-point scale (1 = central; 2 = peripheral; 3 = marginal; 4 = off target), the mean-targeted precision was 1.9 {+-} 0.9. No misplacement of the biopsy canula affecting adjacent structures could be detected. The novel steerable biopsy canula proved to be efficient in the ex-vivo set-up. The chest phantom enabling respiratory motion and the steerable biopsy canula offer a feasible ex-vivo system for evaluating and training CT fluoroscopy-guided lung biopsy adapted to respiratory motion.« less

  9. Percutaneous transfemoral-transseptal implantation of a second-generation CardiAQ mitral valve bioprosthesis: first procedure description and 30-day follow-up.

    PubMed

    Ussia, Gian Paolo; Quadri, Arshad; Cammalleri, Valeria; De Vico, Pasquale; Muscoli, Saverio; Marchei, Massimo; Ruvolo, Giovanni; Sondergaard, Lars; Romeo, Francesco

    2016-02-01

    Transcatheter mitral valve implantation for mitral valve regurgitation is in the very early phase of development because of challenging anatomy and device dimensions. We describe the procedure of a transfemoral-transseptal implantation of the second-generation CardiAQ mitral valve bioprosthesis and 30-day follow-up. The procedure was performed percutaneously, without any left extracorporeal circulatory support. The patient had severe mitral regurgitation with severely depressed ventricular function and other comorbidities. The patient was deemed extreme high risk for conventional cardiac surgery by a multidisciplinary team. The main procedural steps were the creation of an arteriovenous loop with an exchange nitinol wire, and the use of a customised "steerable snare system" to facilitate the catheter delivery system into the mitral annulus. Transoesophageal echocardiography and fluoroscopy were utilised for device positioning and deployment. The mitral valve prosthesis was implanted with mild mitral regurgitation. The postoperative course was uneventful and at 30-day follow-up the patient is in NYHA Class I, with good function of the mitral valve bioprosthesis. This procedure shows that percutaneous transfemoral transcatheter mitral valve implantation is feasible, safe and successful. Further experience is needed to render this procedure clinically available.

  10. Paediatric dose reduction with the introduction of digital fluorography.

    PubMed

    Mooney, R B; McKinstry, J

    2001-01-01

    Fluoroscopy guided examinations in a paediatric X ray department were initially carried out on a unit that used a conventional screen-film combination for spot-films. A new fluoroscopy unit was installed with the facilities of digital fluorography and last image hold. Comparison of equipment performance showed that the dose per image for screen-film and digital fluorography was 3 microGy and 0.4 microGy, respectively. Although the screen-film had superior image quality, the department's radiologist confirmed that digital fluorography provided a diagnostic image. Patient dose measurements showed that introduction of the new unit caused doses to fall by an average of 70%, although fluoroscopy time had not changed significantly. The new unit produced 40% less air kerma during fluoroscopy. The remaining 30% reduction in dose was due to the introduction of digital fluorography and last image hold facilities. It is concluded that the use of digital fluorography can be an effective way of reducing paediatric dose.

  11. Computed tomography angiography-fluoroscopy image fusion allows visceral vessel cannulation without angiography during fenestrated endovascular aneurysm repair.

    PubMed

    Schwein, Adeline; Chinnadurai, Ponraj; Behler, Greg; Lumsden, Alan B; Bismuth, Jean; Bechara, Carlos F

    2018-07-01

    Fenestrated endovascular aneurysm repair (FEVAR) is an evolving technique to treat juxtarenal abdominal aortic aneurysms (AAAs). Catheterization of visceral and renal vessels after the deployment of the fenestrated main body device is often challenging, usually requiring additional fluoroscopy and multiple digital subtraction angiograms. The aim of this study was to assess the clinical utility and accuracy of a computed tomography angiography (CTA)-fluoroscopy image fusion technique in guiding visceral vessel cannulation during FEVAR. Between August 2014 and September 2016, all consecutive patients who underwent FEVAR at our institution using image fusion guidance were included. Preoperative CTA images were fused with intraoperative fluoroscopy after coregistering with non-contrast-enhanced cone beam computed tomography (syngo 3D3D image fusion; Siemens Healthcare, Forchheim, Germany). The ostia of the visceral vessels were electronically marked on CTA images (syngo iGuide Toolbox) and overlaid on live fluoroscopy to guide vessel cannulation after fenestrated device deployment. Clinical utility of image fusion was evaluated by assessing the number of dedicated angiograms required for each visceral or renal vessel cannulation and the use of optimized C-arm angulation. Accuracy of image fusion was evaluated from video recordings by three raters using a binary qualitative assessment scale. A total of 26 patients (17 men; mean age, 73.8 years) underwent FEVAR during the study period for juxtarenal AAA (17), pararenal AAA (6), and thoracoabdominal aortic aneurysm (3). Video recordings of fluoroscopy from 19 cases were available for review and assessment. A total of 46 vessels were cannulated; 38 of 46 (83%) of these vessels were cannulated without angiography but based only on image fusion guidance: 9 of 11 superior mesenteric artery cannulations and 29 of 35 renal artery cannulations. Binary qualitative assessment showed that 90% (36/40) of the virtual ostia overlaid on live fluoroscopy were accurate. Optimized C-arm angulations were achieved in 35% of vessel cannulations (0/9 for superior mesenteric artery cannulation, 12/25 for renal arteries). Preoperative CTA-fluoroscopy image fusion guidance during FEVAR is a valuable and accurate tool that allows visceral and renal vessel cannulation without the need of dedicated angiograms, thus avoiding additional injection of contrast material and radiation exposure. Further refinements, such as accounting for device-induced aortic deformation and automating the image fusion workflow, will bolster this technology toward optimal routine clinical use. Copyright © 2017 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

  12. Performance of mobile digital X-ray fluoroscopy using a novel flat panel detector for intraoperative use.

    PubMed

    Jeong, Chang-Won; Ryu, Jong-Hyun; Joo, Su-Chong; Jun, Hong-Young; Heo, Dong-Woon; Lee, Jinseok; Kim, Kyong-Woo; Yoon, Kwon-Ha

    2015-01-01

    Technologies employing digital X-ray devices are developed for mobile settings. To develop a mobile digital X-ray fluoroscopy (MDF) for intraoperative guidance, using a novel flat panel detector to focus on diagnostics in outpatient clinics, operating and emergency rooms. An MDF for small-scale field diagnostics was configured using an X-ray source and a novel flat panel detector. The imager enabled frame rates reaching 30 fps in full resolution fluoroscopy with maximal running time of 5 minutes. Signal-to-noise (SNR), contrast-to-noise (CNR), and spatial resolution were analyzed. Stray radiation, exposure radiation dose, and effective absorption dose were measured for patients. The system was suitable for small-scale field diagnostics. SNR and CNR were 62.4 and 72.0. Performance at 10% of MTF was 9.6 lp/mm (53 μ m) in the no binned mode. Stray radiation at 100 cm and 150 cm from the source was below 0.2 μ Gy and 0.1 μ Gy. Exposure radiation in radiography and fluoroscopy (5 min) was 10.2 μ Gy and 82.6 mGy. The effective doses during 5-min-long fluoroscopy were 0.26 mSv (wrist), 0.28 mSv (elbow), 0.29 mSv (ankle), and 0.31 mSv (knee). The proposed MDF is suitable for imaging in operating rooms.

  13. Comparison of Fluoroscopy and Computed Tomography for Tracheal Lumen Diameter Measurement and Determination of Intraluminal Stent Size in Healthy Dogs

    PubMed Central

    Williams, Jackie M.; Krebs, Ingar A.; Riedesel, Elizabeth A.; Zhao, Qianqian

    2015-01-01

    Tracheal collapse is a progressive airway disease that can ultimately result in complete airway obstruction. Intraluminal tracheal stents are a minimally invasive and viable treatment for tracheal collapse once the disease becomes refractory to medical management. Intraluminal stent size is chosen based on the maximum measured tracheal diameter during maximum inflation. The purpose of this prospective, cross-sectional study was to compare tracheal lumen diameter measurements and subsequent selected stent size using both fluoroscopy and CT and to evaluate inter- and intraobserver variability of the measurements. Seventeen healthy Beagles were anesthetized and imaged with fluoroscopy and CT with positive pressure ventilation to 20 cm H2O. Fluoroscopic and CT maximum tracheal diameters were measured by 3 readers. Three individual measurements were made at 8 pre-determined tracheal sites for dorsoventral (height) and laterolateral (width) dimensions. Tracheal diameters and stent sizes (based on the maximum tracheal diameter + 10%) were analyzed using a linear mixed model. CT tracheal lumen diameters were larger compared to fluoroscopy at all locations. When comparing modalities, fluoroscopic and CT stent sizes were statistically different. Greater overall variation in tracheal diameter measurement (height or width) existed for fluoroscopy compared to CT, both within and among observers. The greater tracheal diameter and lower measurement variability supported the use of CT for appropriate stent selection to minimize complications in veterinary patients. PMID:26784924

  14. Modeling Cable and Guide Channel Interaction in a High-Strength Cable-Driven Continuum Manipulator

    PubMed Central

    Moses, Matthew S.; Murphy, Ryan J.; Kutzer, Michael D. M.; Armand, Mehran

    2016-01-01

    This paper presents several mechanical models of a high-strength cable-driven dexterous manipulator designed for surgical procedures. A stiffness model is presented that distinguishes between contributions from the cables and the backbone. A physics-based model incorporating cable friction is developed and its predictions are compared with experimental data. The data show that under high tension and high curvature, the shape of the manipulator deviates significantly from a circular arc. However, simple parametric models can fit the shape with good accuracy. The motivating application for this study is to develop a model so that shape can be predicted using easily measured quantities such as tension, so that real-time navigation may be performed, especially in minimally-invasive surgical procedures, while reducing the need for hazardous imaging methods such as fluoroscopy. PMID:27818607

  15. Automatic management system for dose parameters in interventional radiology and cardiology.

    PubMed

    Ten, J I; Fernandez, J M; Vaño, E

    2011-09-01

    The purpose of this work was to develop an automatic management system to archive and analyse the major study parameters and patient doses for fluoroscopy guided procedures performed in cardiology and interventional radiology systems. The X-ray systems used for this trial have the capability to export at the end of the procedure and via e-mail the technical parameters of the study and the patient dose values. An application was developed to query and retrieve from a mail server, all study reports sent by the imaging modality and store them on a Microsoft SQL Server data base. The results from 3538 interventional study reports generated by 7 interventional systems were processed. In the case of some technical parameters and patient doses, alarms were added to receive malfunction alerts so as to immediately take appropriate corrective actions.

  16. Modeling Cable and Guide Channel Interaction in a High-Strength Cable-Driven Continuum Manipulator.

    PubMed

    Moses, Matthew S; Murphy, Ryan J; Kutzer, Michael D M; Armand, Mehran

    2015-12-01

    This paper presents several mechanical models of a high-strength cable-driven dexterous manipulator designed for surgical procedures. A stiffness model is presented that distinguishes between contributions from the cables and the backbone. A physics-based model incorporating cable friction is developed and its predictions are compared with experimental data. The data show that under high tension and high curvature, the shape of the manipulator deviates significantly from a circular arc. However, simple parametric models can fit the shape with good accuracy. The motivating application for this study is to develop a model so that shape can be predicted using easily measured quantities such as tension, so that real-time navigation may be performed, especially in minimally-invasive surgical procedures, while reducing the need for hazardous imaging methods such as fluoroscopy.

  17. [Minimally invasive interventional techniques involving the urogenital tract in dogs and cats].

    PubMed

    Heilmann, R M

    2016-01-01

    Minimally invasive interventional techniques are advancing fast in small animal medicine. These techniques utilize state-of-the-art diagnostic methods, including fluoroscopy, ultrasonography, endoscopy, and laparoscopy. Minimally invasive procedures are particularly attractive in the field of small animal urology because, in the past, treatment options for diseases of the urogenital tract were rather limited or associated with a high rate of complications. Most endourological interventions have a steep learning curve. With the appropriate equipment and practical training some of these procedures can be performed in most veterinary practices. However, most interventions require referral to a specialty clinic. This article summarizes the standard endourological equipment and materials as well as the different endourological interventions performed in dogs and cats with diseases of the kidneys/renal pelves, ureters, or lower urinary tract (urinary bladder and urethra).

  18. Primary percutaneous coronary intervention by magnetic navigation compared with conventional wire technique.

    PubMed

    Patterson, Mark S; Dirksen, Maurits T; Ijsselmuiden, Alexander J; Amoroso, Giovanni; Slagboom, Ton; Laarman, Gerrit-Jan; Schultz, Carl; van Domburg, Ron T; Serruys, Patrick W; Kiemeneij, Ferdinand

    2011-06-01

    Aims Comparison of magnetic guidewire navigation in percutaneous coronary intervention (MPCI) vs. conventional percutaneous coronary intervention (CPCI) for the treatment of acute myocardial infarction. Methods and results We compared 65 sequential patients (mean age 61 ± 15 years) undergoing primary MPCI with those of 405 patients undergoing CPCI (mean age 61 ± 13 years). The major endpoint was contrast media use. Technical success and procedural outcomes were evaluated. Clinical demographics and angiographic characteristics of the two groups were similar, except for fewer patients with previous coronary artery bypass grafting (CABG) and hypertension in the CPCI group and fewer patients with diabetes in the MPCI group. The technical success rate was high in both the MPCI and CPCI groups (95.4 vs. 98%). There was significantly less contrast media usage in the MPCI compared with the CPCI group, median reduction of contrast media of 30 mL with an OR = 0.41 (0.21-0.81). Fluoroscopy times were significantly reduced for MPCI compared with CPCI, median reduction of 7.2 min with an OR = 0.42 (0.20-0.79). Conclusion This comparison indicates the feasibility and non-inferiority of magnetic navigation in performing primary PCI and suggests the possibility of reductions in contrast media use and fluoroscopy time compared with CPCI.

  19. Routine Chest Radiographs After Central Line Insertion: Mandatory Postprocedural Evaluation or Unnecessary Waste of Resources?

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lucey, Brian; Varghese, Jose C.; Haslam, Philip

    1999-09-15

    Purpose: To study the cost and impact on patient management of the routine performance of chest radiographs in patients undergoing imaged-guided central venous catheter insertion. Methods: Six hundred and twenty-one catheters placed in 489 patients over a 42-month period formed the study group. Catheters were placed in the right internal jugular vein (425), left internal jugular vein (133), and subclavian veins (63). At the end of the procedure fluoroscopy was used to assess catheter position and check for complications. A postprocedural chest radiograph was obtained in all patients. Results: Postprocedural chest fluoroscopy showed no evidence of pneumothorax, hemothorax, or mediastinalmore » hematoma. Inappropriate catheter tip position or catheter kinks were noted with 90 catheters. These problems were all corrected while the patient was on the interventional table. Postprocedural chest radiographs showed no complications but proximal catheter tip migration was noted in six of 621 catheters (1%). These latter six catheters required further manipulation. The total technical and related charges for the postprocedural chest radiographs in this series were estimated at Pounds 15,525. Conclusion: Postprocedural chest radiographs after image-guided central venous catheter insertion are not routinely required. A postprocedural chest radiograph can be performed on a case-by-case basis at the discretion of the interventional radiologist.« less

  20. Cervical spinal motion during intubation: efficacy of stabilization maneuvers in the setting of complete segmental instability.

    PubMed

    Lennarson, P J; Smith, D W; Sawin, P D; Todd, M M; Sato, Y; Traynelis, V C

    2001-04-01

    The purpose of this study was to characterize and compare segmental cervical motion during orotracheal intubation in cadavers with and without a complete subaxial injury, as well as to examine the efficacy of commonly used stabilization techniques in limiting that motion. Intubation procedures were performed in 10 fresh human cadavers in which cervical spines were intact and following the creation of a complete C4-5 ligamentous injury. Movement of the cervical spine during direct laryngoscopy and intubation was recorded using video fluoroscopy and examined under the following conditions: 1) without stabilization; 2) with manual in-line cervical immobilization; and 3) with Gardner-Wells traction. Subsequently, segmental angular rotation, subluxation, and distraction at the injured C4-5 level were measured from digitized frames of the recorded video fluoroscopy. After complete C4-5 destabilization, the effects of attempted stabilization on distraction, angulation, and subluxation were analyzed. Immobilization effectively eliminated distraction, and diminished angulation, but increased subluxation. Traction significantly increased distraction, but decreased angular rotation and effectively eliminated subluxation. Orotracheal intubation without stabilization had intermediate results, causing less distraction than traction, less subluxation than immobilization, but increased angulation compared with either intervention. These results are discussed in terms of both statistical and clinical significance and recommendations are made.

  1. Robotic navigation and ablation.

    PubMed

    Malcolme-Lawes, L; Kanagaratnam, P

    2010-12-01

    Robotic technologies have been developed to allow optimal catheter stability and reproducible catheter movements with the aim of achieving contiguous and transmural lesion delivery. Two systems for remote navigation of catheters within the heart have been developed; the first is based on a magnetic navigation system (MNS) Niobe, Stereotaxis, Saint-Louis, Missouri, USA, the second is based on a steerable sheath system (Sensei, Hansen Medical, Mountain View, CA, USA). Both robotic and magnetic navigation systems have proven to be feasible for performing ablation of both simple and complex arrhythmias, particularly atrial fibrillation. Studies to date have shown similar success rates for AF ablation compared to that of manual ablation, with many groups finding a reduction in fluoroscopy times. However, the early learning curve of cases demonstrated longer procedure times, mainly due to additional setup times. With centres performing increasing numbers of robotic ablations and the introduction of a pressure monitoring system, lower power settings and instinctive driving software, complication rates are reducing, and fluoroscopy times have been lower than manual ablation in many studies. As the demand for catheter ablation for arrhythmias such as atrial fibrillation increases and the number of centres performing these ablations increases, the demand for systems which reduce the hand skill requirement and improve the comfort of the operator will also increase.

  2. Fluoroscopy-guided hydrostatic reduction of intussusception in infancy: role of pharmacological premedication.

    PubMed

    Esposito, Francesco; Ambrosio, Concetta; De Fronzo, Simona; Panico, Maria Rita; D'Aprano, Marilena; Giugliano, Anna Marcella; Noviello, Domenico; Oresta, Patrizia

    2015-06-01

    Intussusception is one of the most common causes of paediatric emergency. Fluoroscopy-guided hydrostatic reduction is a common nonoperative management strategy for the treatment of intussusception. The role of pharmacological premedication in increasing the success rate of hydrostatic reduction is still controversial. The purpose of this study was to verify the presence of a possible correlation between pharmacological premedication and the percentage of hydrostatic reduction of intussusception in paediatric patients. This study considered children with a diagnosis of idiopathic intussusception treated at our hospital between January 2007 and June 2013. One group of patients underwent hydrostatic reduction by barium enema without any preliminary therapy. A second group of patients received pharmacological premedication with both a sedative and an anti-oedematous agent before the procedure. A total of 398 patients were treated with barium enema for therapeutic purposes. In the group of patients who received no premedication (n = 254), 165 (65 %) children achieved hydrostatic reduction of the intussusception. Among the patients who received pharmacological premedication prior to barium enema (n = 144), 122 (85 %) children achieved resolution of the intussusception. Our study shows that the use of pharmacological premedication is effective for the reduction of the intussusception, as its limit patient stress, fluoroscopic time and radiation dose.

  3. Temporal response improvement for computed tomography fluoroscopy

    NASA Astrophysics Data System (ADS)

    Hsieh, Jiang

    1997-10-01

    Computed tomography fluoroscopy (CTF) has attracted significant attention recently. This is mainly due to the growing clinical application of CTF in interventional procedures, such as guided biopsy. Although many studies have been conducted for its clinical efficacy, little attention has been paid to the temporal response and the inherent limitations of the CTF system. For example, during a biopsy operation, when needle is inserted at a relatively high speed, the true needle position will not be correctly depicted in the CTF image due to the time delay. This could result in an overshoot or misplacement of the biopsy needle by the operator. In this paper, we first perform a detailed analysis of the temporal response of the CTF by deriving a set of equations to describe the average location of a moving object observed by the CTF system. The accuracy of the equations is verified by computer simulations and experiments. We show that the CT reconstruction process acts as a low pass filter to the motion function. As a result, there is an inherent time delay in the CTF process to the true biopsy needle motion and locations. Based on this study, we propose a generalized underscan weighting scheme which significantly improve the performance of CTF in terms of time lag and delay.

  4. Comparison of cryoablation with 3D mapping versus conventional mapping for the treatment of atrioventricular re-entrant tachycardia and right-sided paraseptal accessory pathways.

    PubMed

    Russo, Mario S; Drago, Fabrizio; Silvetti, Massimo S; Righi, Daniela; Di Mambro, Corrado; Placidi, Silvia; Prosperi, Monica; Ciani, Michele; Naso Onofrio, Maria T; Cannatà, Vittorio

    2016-06-01

    Aim Transcatheter cryoablation is a well-established technique for the treatment of atrioventricular nodal re-entry tachycardia and atrioventricular re-entry tachycardia in children. Fluoroscopy or three-dimensional mapping systems can be used to perform the ablation procedure. The aim of this study was to compare the success rate of cryoablation procedures for the treatment of right septal accessory pathways and atrioventricular nodal re-entry circuits in children using conventional or three-dimensional mapping and to evaluate whether three-dimensional mapping was associated with reduced patient radiation dose compared with traditional mapping. In 2013, 81 children underwent transcatheter cryoablation at our institution, using conventional mapping in 41 children - 32 atrioventricular nodal re-entry tachycardia and nine atrioventricular re-entry tachycardia - and three-dimensional mapping in 40 children - 24 atrioventricular nodal re-entry tachycardia and 16 atrioventricular re-entry tachycardia. Using conventional mapping, the overall success rate was 78.1 and 66.7% in patients with atrioventricular nodal re-entry tachycardia or atrioventricular re-entry tachycardia, respectively. Using three-dimensional mapping, the overall success rate was 91.6 and 75%, respectively (p=ns). The use of three-dimensional mapping was associated with a reduction in cumulative air kerma and cumulative air kerma-area product of 76.4 and 67.3%, respectively (p<0.05). The use of three-dimensional mapping compared with the conventional fluoroscopy-guided method for cryoablation of right septal accessory pathways and atrioventricular nodal re-entry circuits in children was associated with a significant reduction in patient radiation dose without an increase in success rate.

  5. Percutaneous insertion of peritoneal dialysis catheters using ultrasound and fluoroscopic guidance: A single centre experience and review of literature.

    PubMed

    De Boo, Diederick W; Mott, Nigel; Tregaskis, Peter; Quach, Trung; Menahem, Solomon; Walker, Rowan G; Koukounaras, Jim

    2015-12-01

    Various methods of peritoneal dialysis (PD) catheter insertion are available. The purpose of this study was to evaluate a percutaneous insertion technique using ultrasound (US) and fluoroscopy performed under conscious sedation and as day case procedure. Data of 87 percutaneous inserted dialysis catheters were prospectively collected, including patients' age, gender, body mass index, history of previous abdominal surgery and cause of end stage renal failure. Length of hospital stay, early complications and time to first use were also recorded. Institutional review board approval was obtained. A 100% technical success rate was observed. Early complications included bleeding (n = 3), catheter dysfunction (n = 6), exit site infection (n = 1) and exit site leakage (n = 1). All cases of catheter dysfunction and one case of bleeding required surgical revision. Median time of follow-up was 18 months (range 3-35), and median time from insertion to first use was days 14 (1-47). Of the 82 patients who started dialysis, 20 (23%) ceased PD at some stage during follow-up. Most frequently encountered reasons include deteriorating patient cognitive or functional status (n = 5), successful transplant kidney (n = 4) and pleuro-peritoneal fistula (n = 4). Sixty-two (71%) PD catheter insertions were performed as day case. The remaining insertions were performed on patients already admitted to the hospital. Percutaneous insertion of dialysis catheter using US and fluoroscopy is not only safe but can be performed as day case procedure in most patients, even with a medical history of abdominal surgery and/or obesity. © 2015 The Royal Australian and New Zealand College of Radiologists.

  6. The magnetic navigation system allows safety and high efficacy for ablation of arrhythmias.

    PubMed

    Bauernfeind, Tamas; Akca, Ferdi; Schwagten, Bruno; de Groot, Natasja; Van Belle, Yves; Valk, Suzanne; Ujvari, Barbara; Jordaens, Luc; Szili-Torok, Tamas

    2011-07-01

    We aimed to evaluate the safety and long-term efficacy of the magnetic navigation system (MNS) in a large number of patients. The MNS has the potential for improving safety and efficacy based on atraumatic catheter design and superior navigation capabilities. In this study, 610 consecutive patients underwent ablation. Patients were divided into two age- and sex-matched groups. Ablations were performed either using MNS (group MNS, 292) or conventional manual ablation [group manual navigation (MAN), 318]. The following parameters were analysed: acute success rate, fluoroscopy time, procedure time, complications [major: pericardial tamponade, permanent atrioventricular (AV) block, major bleeding, and death; minor: minor bleeding and temporary AV block]. Recurrence rate was assessed during follow-up (15±9.5 months). Subgroup analysis was performed for the following groups: atrial fibrillation, isthmus dependent and atypical atrial flutter, atrial tachycardia, AV nodal re-entrant tachycardia, circus movement tachycardia, and ventricular tachycardia (VT). Magnetic navigation system was associated with less major complications (0.34 vs. 3.2%, P=0.01). The total numbers of complications were lower in group MNS (4.5 vs. 10%, P=0.005). Magnetic navigation system was equally effective as MAN in acute success rate for overall groups (92 vs. 94%, P=ns). Magnetic navigation system was more successful for VTs (93 vs. 72%, P<0.05). Less fluoroscopy was used in group MNS (30±20 vs. 35±25 min, P<0.01). There were no differences in procedure times and recurrence rates for the overall groups (168±67 vs. 159±75 min, P=ns; 14 vs. 11%, P=ns; respectively). Our data suggest that the use of MNS improves safety without compromising efficiency of ablations. Magnetic navigation system is more effective than manual ablation for VTs.

  7. Rigorous accuracy assessment for 3D reconstruction using time-series Dual Fluoroscopy (DF) image pairs

    NASA Astrophysics Data System (ADS)

    Al-Durgham, Kaleel; Lichti, Derek D.; Kuntze, Gregor; Ronsky, Janet

    2017-06-01

    High-speed biplanar videoradiography, or clinically referred to as dual fluoroscopy (DF), imaging systems are being used increasingly for skeletal kinematics analysis. Typically, a DF system comprises two X-ray sources, two image intensifiers and two high-speed video cameras. The combination of these elements provides time-series image pairs of articulating bones of a joint, which permits the measurement of bony rotation and translation in 3D at high temporal resolution (e.g., 120-250 Hz). Assessment of the accuracy of 3D measurements derived from DF imaging has been the subject of recent research efforts by several groups, however with methodological limitations. This paper presents a novel and simple accuracy assessment procedure based on using precise photogrammetric tools. We address the fundamental photogrammetry principles for the accuracy evaluation of an imaging system. Bundle adjustment with selfcalibration is used for the estimation of the system parameters. The bundle adjustment calibration uses an appropriate sensor model and applies free-network constraints and relative orientation stability constraints for a precise estimation of the system parameters. A photogrammetric intersection of time-series image pairs is used for the 3D reconstruction of a rotating planar object. A point-based registration method is used to combine the 3D coordinates from the intersection and independently surveyed coordinates. The final DF accuracy measure is reported as the distance between 3D coordinates from image intersection and the independently surveyed coordinates. The accuracy assessment procedure is designed to evaluate the accuracy over the full DF image format and a wide range of object rotation. Experiment of reconstruction of a rotating planar object reported an average positional error of 0.44 +/- 0.2 mm in the derived 3D coordinates (minimum 0.05 and maximum 1.2 mm).

  8. Electroanatomic Mapping-Guided Catheter Ablation of Supraventricular Tachycardia in Children with Ebstein's Anomaly.

    PubMed

    Ergul, Yakup; Koca, Serhat; Akdeniz, Celal; Tuzcu, Volkan

    2018-06-07

    In Ebstein's anomaly (EA), tachycardia substrates are complex, and accessory pathway (AP) ablations are often challenging. This study demonstrates the utility of the EnSite Velocity system (St. Jude Medical, St Paul, MN) in the catheter ablation of supraventricular tachycardia in children with EA. Twenty patients [Female/Male = 8/12, median age 11.5 years (2.6-18)] with EA who underwent catheter ablation guided by the EnSite Velocity system between December 2011 and December 2016 were retrospectively evaluated. Five patients had severe EA, and two of them were at Fontan palliation pathway. The most common indications for ablations were palpitations/syncope and treatment-resistant arrhythmias. Thirty-one tachycardia substrate foci (21 manifest AP, 2 concealed AP, 4 Mahaim AP, 3 focal atrial tachycardias, and 1 typical atrioventricular nodal reentrant tachycardia) were detected in 20 patients. There were multiple tachycardia substrates in 11 patients (55%). The patient-based acute procedure success rate was 19/20 (95%), and the tachycardia-based success rate was 30/31 (97%). The mean procedure time was 170 ± 43 min (90-265). Fluoroscopy was not used in 15 (75%) patients. The mean fluoroscopy time in the remaining five patients was 3.6 ± 2.9 min (0.7-7.8). During a mean follow-up of 35.1 ± 20.3 months (6-60), tachycardia recurred in four patients (4/19, 21%). No complications were seen. Catheter ablation of arrhythmias can be performed effectively and safely in pediatric EA patients by using a limited fluoroscopic approach with the help of electroanatomical mapping systems. However, the rate of tachycardia recurrence at follow-up remains high.

  9. Transcatheter closure of patent ductus arteriosus using the AMPLATZER™ duct occluder II (ADO II).

    PubMed

    Gruenstein, Daniel H; Ebeid, Makram; Radtke, Wolfgang; Moore, Phillip; Holzer, Ralf; Justino, Henri

    2017-05-01

    The study purpose is to evaluate the safety and efficacy of the ADO II device for closure of patent ductus arteriosus (PDA) in children. Transcatheter treatment of PDA has been evolving for 40+ years and is the treatment of choice. The AMPLATZER™ Duct Occluder (ADO) device was developed for larger diameter ducts and is not ideal in all PDAs. ADO II was developed for small to moderate-sized ducts. This is a single-arm, multicenter study evaluating safety and efficacy of the ADO II device. Patients <18 years were screened for a PDA ≤5.5 mm in diameter and 3-12 mm in length. Right and left heart catheterization was performed, and hemodynamic data were obtained at the time of implant. The diameter of the left pulmonary artery (LPA) and descending aorta, and the presence of any pre-existing pressure gradients across the LPA or aortic arch were assessed at baseline and 6 months post-implant. A total of 192 patients were enrolled. The median implant time was 74 min. Median fluoroscopy time was 12 min. A retrograde (aortic) approach was used in 33% of procedures and demonstrated a statistically significant reduction in fluoroscopy time (P value = 0.0018) compared to an antegrade approach. The device was successfully implanted in 93% of patients, with complete closure in 98% of successful implantations. In this prospective study, the ADO II was safe and effective for closure of small to moderate PDAs. Implantation is simple and the ability for retrograde aortic delivery reduces procedure-related radiation exposure. © 2017 Wiley Periodicals, Inc. © 2017 Wiley Periodicals, Inc.

  10. Temporal trends in safety and complication rates of catheter ablation for atrial fibrillation.

    PubMed

    Muthalaly, Rahul G; John, Roy M; Schaeffer, Benjamin; Tanigawa, Shinichi; Nakamura, Tomofumi; Kapur, Sunil; Zei, Paul C; Epstein, Laurence M; Tedrow, Usha B; Michaud, Gregory F; Stevenson, William G; Koplan, Bruce A

    2018-06-01

    Atrial fibrillation (AF) ablation is increasingly common, but is associated with potential major complications. Technology, experience, and protocols have evolved significantly in recent times, and may have impacted procedural safety. We sought to compare AF ablation safety profiles, including complication rates and fluoroscopy times in a "modern" versus "historical" cohort. We evaluated consecutive patients undergoing AF ablation from a modern cohort (MC) from 2014 to 2015 and a historic cohort (HC) from 2009 to 2011 for complications. Major complications were categorized according to Heart Rhythm Society guidelines. We included 1,425 patients, 726 in the HC and 699 in the MC. The MC was older, had more OSA and less valvular AF. Fifty-two (3.5%) procedures suffered major complications across the cohorts, with significantly fewer in the MC (5.0% vs. 2.3%, P  =  0.007). The largest reductions were seen in vascular, hemorrhagic, ischemic stroke, and perforation/tamponade related complications. Periprocedural antiplatelets drugs (aHR 2.1 [95 CI 1.1-3.9], P  =  0.02) and force-sensing catheters (aHR 0.4 [95 CI 0.2-0.9], P  =  0.03) were independently related to major complication rates. Direct oral anticoagulants and uninterrupted anticoagulation were not associated with complications. There was a decrease in both fluoroscopy (-17.4 minutes [95 CI 19.2-15.6], P < 0.0001) and radiofrequency ablation times (-561 seconds [95CI -750 to -371], P < 0.0001). The safety profile of AF ablation has improved significantly in less than a decade. © 2018 Wiley Periodicals, Inc.

  11. Robotically Driven CT-guided Needle Insertion: Preliminary Results in Phantom and Animal Experiments.

    PubMed

    Hiraki, Takao; Kamegawa, Tetsushi; Matsuno, Takayuki; Sakurai, Jun; Kirita, Yasuzo; Matsuura, Ryutaro; Yamaguchi, Takuya; Sasaki, Takanori; Mitsuhashi, Toshiharu; Komaki, Toshiyuki; Masaoka, Yoshihisa; Matsui, Yusuke; Fujiwara, Hiroyasu; Iguchi, Toshihiro; Gobara, Hideo; Kanazawa, Susumu

    2017-11-01

    Purpose To evaluate the accuracy of the remote-controlled robotic computed tomography (CT)-guided needle insertion in phantom and animal experiments. Materials and Methods In a phantom experiment, 18 robotic and manual insertions each were performed with 19-gauge needles by using CT fluoroscopic guidance for the evaluation of the equivalence of accuracy of insertion between the two groups with a 1.0-mm margin. Needle insertion time, CT fluoroscopy time, and radiation exposure were compared by using the Student t test. The animal experiments were approved by the institutional animal care and use committee. In the animal experiment, five robotic insertions each were attempted toward targets in the liver, kidneys, lungs, and hip muscle of three swine by using 19-gauge or 17-gauge needles and by using conventional CT guidance. The feasibility, safety, and accuracy of robotic insertion were evaluated. Results The mean accuracies of robotic and manual insertion in phantoms were 1.6 and 1.4 mm, respectively. The 95% confidence interval of the mean difference was -0.3 to 0.6 mm. There were no significant differences in needle insertion time, CT fluoroscopy time, or radiation exposure to the phantom between the two methods. Effective dose to the physician during robotic insertion was always 0 μSv, while that during manual insertion was 5.7 μSv on average (P < .001). Robotic insertion was feasible in the animals, with an overall mean accuracy of 3.2 mm and three minor procedure-related complications. Conclusion Robotic insertion exhibited equivalent accuracy as manual insertion in phantoms, without radiation exposure to the physician. It was also found to be accurate in an in vivo procedure in animals. © RSNA, 2017 Online supplemental material is available for this article.

  12. Lower Learning Difficulty and Fluoroscopy Reduction of Transforaminal Percutaneous Endoscopic Lumbar Discectomy with an Accurate Preoperative Location Method.

    PubMed

    Fan, Guoxin; Gu, Xin; Liu, Yifan; Wu, Xinbo; Zhang, Hailong; Gu, Guangfei; Guan, Xiaofei; He, Shisheng

    2016-01-01

    Transforaminal percutaneous endoscopic lumbar discectomy (tPELD) poses great challenges for junior surgeons. Beginners often require repeated attempts using fluoroscopy causing more punctures, which may significantly undermine their confidence and increase the radiation exposure to medical staff and patients. Moreover, the impact of an accurate location on the learning curve of tPELD has not been defined. The study aimed to investigate the impact of an accurate preoperative location method on learning difficulty and fluoroscopy time of tPELD. Retrospective evaluation. Patients receiving tPELD by one surgeon with a novel accurate preoperative location method were regarded as Group A, and those receiving tPELD by another surgeon with a conventional fluoroscopy method were regarded as Group B. From January 2012 to August 2014, we retrospectively reviewed the first 80 tPELD cases conducted by 2 junior surgeons. The operation time, fluoroscopy times, preoperative location time, and puncture-channel time were thoroughly analyzed. The operation time of the first 20 patients were 99.75 ± 10.38 minutes in Group A and 115.7 ± 16.46 minutes in Group B, while the operation time of all 80 patients was 88.36 ± 11.56 minutes in Group A and 98.26 ± 14.90 minutes in Group B. Significant differences were detected in operation time between the 2 groups, both for the first 20 patients and total 80 patients (P < 0.05). The fluoroscopy times were 26.78 ± 4.17 in Group A and 33.98 ± 2.69 in Group B (P < 0.001). The preoperative location time was 3.43 ± 0.61 minutes in Group A and 5.59 ± 1.46 minutes in Group B (P < 0.001). The puncture-channel time was 27.20 ± 4.49 minutes in Group A and 34.64 ± 8.35 minutes in Group B (P < 0.001). There was a moderate correlation between preoperative location time and puncture-channel time (r = 0.408, P < 0.001), and a moderate correlation between preoperative location time and fluoroscopy times (r = 0.441, P < 0.001). Mild correlations were also observed between preoperative location time and operation time (r = 0.270, P = 0.001). There were no significant differences in preoperative back visual analogue scale (VAS) score, postoperative back VAS, preoperative leg VAS, postoperative leg VAS, preoperative Japanese Orthopaedic Association (JOA) score, postoperative JOA, preoperative Oswestry disability score (ODI), or postoperative ODI (P > 0.05). However, significant differences were all detected between preoperative abovementioned scores and postoperative scores (P < 0.05). Moreover, there was no significant differences in Macnab satisfaction between the 2 groups (P = 0.179). There were 2 patients with recurrence in Group A and 3 patients in Group B. Twelve patients with postoperative disc remnants were identified in Group A and 9 patients in Group B. No significant difference was identified between the 2 groups (P = 0.718). The preoperative lumbar location method is just a tiny step in tPELD, junior surgeons still need to focus on their subjective feelings during punctures and accumulating their experience in endoscopic discectomy. The accurate preoperative location method lowered the learning difficulty and reduced the fluoroscopy time of tPELD, which was also associated with lower preoperative location time and puncture-channel time. Key words: Learning difficulty, fluoroscopy reduction, transforamimal percutaneous endoscopic lumbar discectomy, preoperative locationLearning difficulty, fluoroscopy reduction, transforamimal percutaneous endoscopic lumbar discectomy, preoperative location.

  13. Use of the Uro Dyna-CT in endourology - the new frontier.

    PubMed

    Vicentini, Fabio C; Botelho, Luiz A A; Braz, José L M; Almeida, Ernane S; Hisano, Marcelo

    2017-01-01

    We describe the use of the Uro Dyna-CT, an imaging system used in the operating room that produces real-time three-dimensional (3D) imaging and cross-sectional image reconstructions similar to an intraoperative computerized tomography, during a percutaneous nephrolithotomy and a contralateral flexible ureteroscopy in a complete supine position. A 65 year-old female patient had an incomplete calyceal staghorn stone in the right kidney and a 10mm in the left one. The procedure was uneventful and the intraoperative use of the Uro Dyna-CT identified 2 residual stones that were not found by digital fluoroscopy and flexible nephroscopy at the end of surgery, helping us to render the patient stone-free in one procedure, which was confirmed by a postoperative CT scan. Prospective studies will define the real role of the Uro Dyna-CT for endourological procedures, but its use seems to be a very promising tool for improving stone free rates and decreasing auxiliary procedures, especially for complex cases. Copyright® by the International Brazilian Journal of Urology.

  14. Assessment of eye lens doses for workers during interventional radiology procedures.

    PubMed

    Urboniene, A; Sadzeviciene, E; Ziliukas, J

    2015-07-01

    The assessment of eye lens doses for workers during interventional radiology (IR) procedures was performed using a new eye lens dosemeter. In parallel, the results of routine individual monitoring were analysed and compared with the results obtained from measurements with a new eye lens dosemeter. The eye lens doses were assessed using Hp(3) measured at the level of the eyes and were compared with Hp(10) measured with the whole-body dosemeter above the lead collar. The information about use of protective measures, the number of performed interventional procedures per month and their fluoroscopy time was also collected. The assessment of doses to the lens of the eye was done for 50 IR workers at 9 Lithuanian hospitals for the period of 2012-2013. If the use of lead glasses is not taken into account, the estimated maximum annual dose equivalent to the lens of the eye was 82 mSv. © The Author 2015. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  15. Reduction in radiation dose for atrial fibrillation ablation over time: A 12-year single-center experience of 2344 patients.

    PubMed

    Voskoboinik, Aleksandr; Kalman, Elana S; Savicky, Yonatan; Sparks, Paul B; Morton, Joseph B; Lee, Geoffrey; Kistler, Peter M; Kalman, Jonathan M

    2017-06-01

    Pulmonary vein isolation (PVI) is a well-established treatment of atrial fibrillation (AF), with contact force (CF)-sensing catheters joining 3-dimensional mapping systems and image integration as technological advancements over the last decade. The purpose of this study was to analyze trends in radiation exposure for AF ablation over the last 12 years at our center. We reviewed prospectively collected data of 2344 consecutive PVI procedures for either paroxysmal or persistent AF between January 2004 and December 2015. During this period, all cases used 3-dimensional mapping systems, with 8 software and 2 hardware upgrades. Primary endpoints were fluoroscopy time, absorbed dose (Air Kerma in mGy), and effective dose (mSv). In total, 1914 patients underwent initial PVI, and 430 patients underwent redo PVI using radiofrequency energy. Fluoroscopy time, and absorbed and effective doses significantly and progressively decreased over the 12-year period for initial PVI as follows: 2004-2006: 61 ± 27 minutes; 2007-2009: 46 ± 14 minutes, 1365 ± 1369 mGy, 11.3 ± 12.5 mSv; 2010-2012: 31 ± 11, 464 ± 339 mGy, 9.0 ± 10.4 mSv; and 2013-2015: 17 ± 9 minutes, 304 ± 758 mGy, 5.5 ± 6.7 mSv. CF-sensing catheters were used for 357/508 PVI only cases between 2014 and 2015. Fluoroscopy times (11 ± 5 vs 21 ± 8 minutes; P <.001) and absorbed dose (200 ± 524 vs 470 ± 1326 mGy; P = .004) were significantly shorter with this catheter. Radiation exposure has dramatically decreased over the last decade for PVI and is related to operator experience, annual case volume, technology evolution, and more recently CF-sensing catheters. This has significant implications for both patient and operator long-term risk. Copyright © 2017 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

  16. Assessing patient dose in interventional fluoroscopy using patient-dependent hybrid phantoms

    NASA Astrophysics Data System (ADS)

    Johnson, Perry Barnett

    Interventional fluoroscopy uses ionizing radiation to guide small instruments through blood vessels or other body pathways to sites of clinical interest. The technique represents a tremendous advantage over invasive surgical procedures, as it requires only a small incision, thus reducing the risk of infection and providing for shorter recovery times. The growing use and increasing complexity of interventional procedures, however, has resulted in public health concerns regarding radiation exposures, particularly with respect to localized skin dose. Tracking and documenting patient-specific skin and internal organ dose has been specifically identified for interventional fluoroscopy where extended irradiation times, multiple projections, and repeat procedures can lead to some of the largest doses encountered in radiology. Furthermore, inprocedure knowledge of localized skin doses can be of significant clinical importance to managing patient risk and in training radiology residents. In this dissertation, a framework is presented for monitoring the radiation dose delivered to patients undergoing interventional procedures. The framework is built around two key points, developing better anthropomorphic models, and designing clinically relevant software systems for dose estimation. To begin, a library of 50 hybrid patient-dependent computational phantoms was developed based on the UF hybrid male and female reference phantoms. These phantoms represent a different type of anthropomorphic model whereby anthropometric parameters from an individual patient are used during phantom selection. The patient-dependent library was first validated and then used in two patient-phantom matching studies focused on cumulative organ and local skin dose. In terms of organ dose, patient-phantom matching was shown most beneficial for estimating the dose to large patients where error associated with soft tissue attenuation differences could be minimized. For small patients, inherent difference in organ size and location limited the effectiveness of matching. For skin dose, patient-phantom matching was found most beneficial for estimating the dose during lateral and anterior-posterior projections. Patient-sculpting of the patient.s outer body contour was also investigated for use during skin dose estimation and highlighted as a substantial step towards better patient-specificity. In order to utilize the models for actual patient dosimetry, two programs were developed based on the newly released Radiation Dose Structured Report (RDSR). The first program allows for the visualization of skin dose by translating the reference point air kerma to the location of the patient.s skin characterized by a computational model. The program represents an innovative tool that can be used by the interventional physician to modify behavior when clinically appropriate. The second program operates by automatically generating an input file from the RDSR which can then be run within a Monte Carlo based radiation transport code. The program has great potential for initiating and promoting the concept of 'cloud dosimetry', where patient-specific radiation transport is performed off-site and returned via the internet. Both programs are non-proprietary and transferable, and also incorporate the most advanced computational phantoms developed to date. Using the tools developed in this work, there exist a tangible opportunity to improve patient care with the end goal being a better understanding of the risk/benefit relationship that accompanies the medical use of ionizing radiation.

  17. Balloon dilatation of nasopharyngeal stenosis in a dog.

    PubMed

    Berent, Allyson C; Kinns, Jennifer; Weisse, Chick

    2006-08-01

    A dog was examined because of a 6-month history of upper airway stridor that began after postoperative regurgitation of gastric contents. Constant stridor was evident during inspiration and expiration, although it was worse during inspiration. The stridor was no longer evident when the dog's mouth was manually held open. Computed tomography, rhinoscopy, and fluoroscopy were used to confirm a diagnosis of nasopharyngeal stenosis. The dog was anesthetized, and balloon dilatation of the stenosis was performed. Prednisone was prescribed for 4 weeks after the procedure to decrease fibrous tissue formation. Although the dog was initially improved, signs recurred 3.5 weeks later, and balloon dilatation was repeated. This time, however, triamcinolone was injected into the area of stenosis at the end of the dilatation procedure. Two months later, although the dog did not have clinical signs of stridor, a third dilatation procedure was performed because mild stenosis was seen on follow-up computed tomographic images; again, triamcinolone was injected into the area of stenosis at the end of the dilatation procedure. Three and 6 months after the third dilatation procedure, the dog reportedly was clinically normal. Findings suggest that balloon dilatation may be an effective treatment for nasopharyngeal stenosis in dogs.

  18. The evaluation of radiologic methods for access guidance in percutaneous nephrolithotomy: a systematic review of the literature.

    PubMed

    Breda, Alberto; Territo, Angelo; Scoffone, Cesare; Seitz, Christian; Knoll, Thomas; Herrmann, Thomas; Brehmer, Mariannhe; Osther, Palle J S; Liatsikos, Evangelos

    2017-11-12

    Percutaneous nephrolithotomy (PNL) is the treatment of choice for larger and complex renal calculi. First step in performing PNL is to obtain access to the renal cavity using either fluoroscopy or ultrasound (US) guidance or a combination of both. Which guiding method to choose is controversial? A systematic review of the literature was performed comparing image guidance modalities for obtaining access in PNL. Evidence acquisition and synthesis: A PubMed, Scopus and Cochrane search for peer-reviewed studies was performed using the keywords "ultrasound" AND "fluoroscopy" AND "Percutaneous nephrolithotomy". Eligible articles were reviewed according to PRISMA criteria. Two hundred and forty records were identified using the keywords. Of these twelve studies were considered relevant. US guidance seems to be associated with a slightly lower complication rate, which may be related to fewer puncture attempts needed for obtaining access and to better peri-renal organ visualization. On the other hand, US-guidance alone needs the adjunct of fluoroscopy in a significant number of cases for achieving access. Stone free rate (SFR) was comparable between groups. Using US for renal access unequivocally reduces radiation exposure. Current evidence indicates that both fluoroscopy and US guidance may be successfully used for obtaining percutaneous renal access. Combining the image-guiding modalities - US and fluoroscopy - seems to increase outcome in PNL both with regard to success in achieving access and reducing complications. Furthermore, including US in the access strategy of PNL reduces radiation exposure to surgeon and staff as well as patients.

  19. Semi-automated intra-operative fluoroscopy guidance for osteotomy and external-fixator.

    PubMed

    Lin, Hong; Samchukov, Mikhail L; Birch, John G; Cherkashin, Alexander

    2006-01-01

    This paper outlines a semi-automated intra-operative fluoroscopy guidance and monitoring approach for osteotomy and external-fixator application in orthopedic surgery. Intra-operative Guidance module is one component of the "LegPerfect Suite" developed for assisting the surgical correction of lower extremity angular deformity. The Intra-operative Guidance module utilizes information from the preoperative surgical planning module as a guideline to overlay (register) its bone outline semi-automatically with the bone edge from the real-time fluoroscopic C-Arm X-Ray image in the operating room. In the registration process, scaling factor is obtained automatically through matching a fiducial template in the fluoroscopic image and a marker in the module. A triangle metal plate, placed on the operating table is used as fiducial template. The area of template image within the viewing area of the fluoroscopy machine is obtained by the image processing techniques such as edge detection and Hough transformation to extract the template from other objects in the fluoroscopy image. The area of fiducial template from fluoroscopic image is then compared with the area of the marker from the planning so as to obtain the scaling factor. After the scaling factor is obtained, the user can use simple operations by mouse to shift and rotate the preoperative planning to overlay the bone outline from planning with the bone edge from fluoroscopy image. In this way osteotomy levels and external fixator positioning on the limb can guided by the computerized preoperative plan.

  20. A Dynamic Image Quality Evaluation of Videofluoroscopy Images: Considerations for Telepractice Applications.

    PubMed

    Burns, Clare L; Keir, Benjamin; Ward, Elizabeth C; Hill, Anne J; Farrell, Anna; Phillips, Nick; Porter, Linda

    2015-08-01

    High-quality fluoroscopy images are required for accurate interpretation of videofluoroscopic swallow studies (VFSS) by speech pathologists and radiologists. Consequently, integral to developing any system to conduct VFSS remotely via telepractice is ensuring that the quality of the VFSS images transferred via the telepractice system is optimized. This study evaluates the extent of change observed in image quality when videofluoroscopic images are transmitted from a digital fluoroscopy system to (a) current clinical equipment (KayPentax Digital Swallowing Workstation, and b) four different telepractice system configurations. The telepractice system configurations consisted of either a local C20 or C60 Cisco TelePresence System (codec unit) connected to the digital fluoroscopy system and linked to a second remote C20 or C60 Cisco TelePresence System via a network running at speeds of either 2, 4 or 6 megabits per second (Mbit/s). Image quality was tested using the NEMA XR 21 Phantom, and results demonstrated some loss in spatial resolution, low contrast detectability and temporal resolution for all transferred images when compared to the fluoroscopy source. When using higher capacity codec units and/or the highest bandwidths to support data transmission, image quality transmitted through the telepractice system was found to be comparable if not better than the current clinical system. This study confirms that telepractice systems can be designed to support fluoroscopy image transfer and highlights important considerations when developing telepractice systems for VFSS analysis to ensure high-quality radiological image reproduction.

  1. Accuracy in identifying the elbow rotation axis on simulated fluoroscopic images using a new anatomical landmark.

    PubMed

    Wiggers, J K; Snijders, R M; Dobbe, J G G; Streekstra, G J; den Hartog, D; Schep, N W L

    2017-11-01

    External fixation of the elbow requires identification of the elbow rotation axis, but the accuracy of traditional landmarks (capitellum and trochlea) on fluoroscopy is limited. The relative distance (RD) of the humerus may be helpful as additional landmark. The first aim of this study was to determine the optimal RD that corresponds to an on-axis lateral image of the elbow. The second aim was to assess whether the use of the optimal RD improves the surgical accuracy to identify the elbow rotation axis on fluoroscopy. CT scans of elbows from five volunteers were used to simulate fluoroscopy; the actual rotation axis was calculated with CT-based flexion-extension analysis. First, three observers measured the optimal RD on simulated fluoroscopy. The RD is defined as the distance between the dorsal part of the humerus and the projection of the posteromedial cortex of the distal humerus, divided by the anteroposterior diameter of the humerus. Second, eight trauma surgeons assessed the elbow rotation axis on simulated fluoroscopy. In a preteaching session, surgeons used traditional landmarks. The surgeons were then instructed how to use the optimal RD as additional landmark in a postteaching session. The deviation from the actual rotation axis was expressed as rotational and translational error (±SD). Measurement of the RD was robust and easily reproducible; the optimal RD was 45%. The surgeons identified the elbow rotation axis with a mean rotational error decreasing from 7.6° ± 3.4° to 6.7° ± 3.3° after teaching how to use the RD. The mean translational error decreased from 4.2 ± 2.0 to 3.7 ± 2.0 mm after teaching. The humeral RD as additional landmark yielded small but relevant improvements. Although fluoroscopy-based external fixator alignment to the elbow remains prone to error, it is recommended to use the RD as additional landmark.

  2. Flexible robotics with electromagnetic tracking improves safety and efficiency during in vitro endovascular navigation.

    PubMed

    Schwein, Adeline; Kramer, Ben; Chinnadurai, Ponraj; Walker, Sean; O'Malley, Marcia; Lumsden, Alan; Bismuth, Jean

    2017-02-01

    One limitation of the use of robotic catheters is the lack of real-time three-dimensional (3D) localization and position updating: they are still navigated based on two-dimensional (2D) X-ray fluoroscopic projection images. Our goal was to evaluate whether incorporating an electromagnetic (EM) sensor on a robotic catheter tip could improve endovascular navigation. Six users were tasked to navigate using a robotic catheter with incorporated EM sensors in an aortic aneurysm phantom. All users cannulated two anatomic targets (left renal artery and posterior "gate") using four visualization modes: (1) standard fluoroscopy mode (control), (2) 2D fluoroscopy mode showing real-time virtual catheter orientation from EM tracking, (3) 3D model of the phantom with anteroposterior and endoluminal view, and (4) 3D model with anteroposterior and lateral view. Standard X-ray fluoroscopy was always available. Cannulation and fluoroscopy times were noted for every mode. 3D positions of the EM tip sensor were recorded at 4 Hz to establish kinematic metrics. The EM sensor-incorporated catheter navigated as expected according to all users. The success rate for cannulation was 100%. For the posterior gate target, mean cannulation times in minutes:seconds were 8:12, 4:19, 4:29, and 3:09, respectively, for modes 1, 2, 3 and 4 (P = .013), and mean fluoroscopy times were 274, 20, 29, and 2 seconds, respectively (P = .001). 3D path lengths, spectral arc length, root mean dimensionless jerk, and number of submovements were significantly improved when EM tracking was used (P < .05), showing higher quality of catheter movement with EM navigation. The EM tracked robotic catheter allowed better real-time 3D orientation, facilitating navigation, with a reduction in cannulation and fluoroscopy times and improvement of motion consistency and efficiency. Copyright © 2016 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

  3. Right heart catheterization using metallic guidewires and low SAR cardiovascular magnetic resonance fluoroscopy at 1.5 Tesla: first in human experience.

    PubMed

    Campbell-Washburn, Adrienne E; Rogers, Toby; Stine, Annette M; Khan, Jaffar M; Ramasawmy, Rajiv; Schenke, William H; McGuirt, Delaney R; Mazal, Jonathan R; Grant, Laurie P; Grant, Elena K; Herzka, Daniel A; Lederman, Robert J

    2018-06-21

    Cardiovascular magnetic resonance (CMR) fluoroscopy allows for simultaneous measurement of cardiac function, flow and chamber pressure during diagnostic heart catheterization. To date, commercial metallic guidewires were considered contraindicated during CMR fluoroscopy due to concerns over radiofrequency (RF)-induced heating. The inability to use metallic guidewires hampers catheter navigation in patients with challenging anatomy. Here we use low specific absorption rate (SAR) imaging from gradient echo spiral acquisitions and a commercial nitinol guidewire for CMR fluoroscopy right heart catheterization in patients. The low-SAR imaging protocol used a reduced flip angle gradient echo acquisition (10° vs 45°) and a longer repetition time (TR) spiral readout (10 ms vs 2.98 ms). Temperature was measured in vitro in the ASTM 2182 gel phantom and post-mortem animal experiments to ensure freedom from heating with the selected guidewire (150 cm × 0.035″ angled-tip nitinol Terumo Glidewire). Seven patients underwent CMR fluoroscopy catheterization. Time to enter each chamber (superior vena cava, main pulmonary artery, and each branch pulmonary artery) was recorded and device visibility and confidence in catheter and guidewire position were scored on a Likert-type scale. Negligible heating (< 0.07°C) was observed under all in vitro conditions using this guidewire and imaging approach. In patients, chamber entry was successful in 100% of attempts with a guidewire compared to 94% without a guidewire, with failures to reach the branch pulmonary arteries. Time-to-enter each chamber was similar (p=NS) for  the two approaches. The guidewire imparted useful catheter shaft conspicuity and enabled interactive modification of catheter shaft stiffness, however, the guidewire tip visibility was poor. Under specific conditions, trained operators can apply low-SAR imaging and using a specific fully-insulated metallic nitinol guidewire (150 cm × 0.035" Terumo Glidewire) to augment clinical CMR fluoroscopy right heart catheterization. Clinicaltrials.gov NCT03152773 , registered May 15, 2017.

  4. [Risk management in regional anesthesia: preface and comments].

    PubMed

    Okuda, Yasuhisa

    2011-11-01

    The benefits of regional anesthesia for surgical procedures, when compared with general anesthesia and/or systemic analgesia, include improved postoperative analgesia, an associated decrease in postoperative pain medication use, decreased nausea and vomiting, and quicker recovery and discharge from the hospital. Neurologic complications associated with regional anesthesia are extremely rare. Although rare, these complications may be reduced with new regional techniques such as the use of ultrasound or fluoroscopy, but further detailed research is needed. In regional anesthesia, rare but serious complications make it necessary to always consider the risk-benefit ratio. The articles discuss these issues and give advice on its effective and safe conduct.

  5. Image Fusion and 3D Roadmapping in Endovascular Surgery.

    PubMed

    Jones, Douglas W; Stangenberg, Lars; Swerdlow, Nicholas J; Alef, Matthew; Lo, Ruby; Shuja, Fahad; Schermerhorn, Marc L

    2018-05-21

    Practitioners of endovascular surgery have historically utilized two-dimensional (2D) intraoperative fluoroscopic imaging, with intra-vascular contrast opacification, to treat complex three-dimensional (3D) pathology. Recently, major technical developments in intraoperative imaging have made image fusion techniques possible: the creation of a 3D patient-specific vascular roadmap based on preoperative imaging which aligns with intraoperative fluoroscopy, with many potential benefits. First, a 3D model is segmented from preoperative imaging, typically a CT scan. The model is then used to plan for the procedure, with placement of specific markers and storing of C-arm angles that will be used for intra-operative guidance. At the time of the procedure, an intraoperative cone-beam CT is performed and the 3D model is registered to the patient's on-table anatomy. Finally, the system is used for live guidance where the 3D model is codisplayed overlying fluoroscopic images. Copyright © 2018. Published by Elsevier Inc.

  6. Correlative anatomy for the electrophysiologist: ablation for atrial fibrillation. Part II: regional anatomy of the atria and relevance to damage of adjacent structures during AF ablation.

    PubMed

    Macedo, Paula G; Kapa, Suraj; Mears, Jennifer A; Fratianni, Amy; Asirvatham, Samuel J

    2010-07-01

    Ablation procedures for atrial fibrillation have become an established and increasingly used option for managing patients with symptomatic arrhythmia. The anatomic structures relevant to the pathogenesis of atrial fibrillation and ablation procedures are varied and include the pulmonary veins, other thoracic veins, the left atrial myocardium, and autonomic ganglia. Exact regional anatomic knowledge of these structures is essential to allow correlation with fluoroscopy and electrograms and, importantly, to avoid complications from damage of adjacent structures within the chest. We present this information as a series of 2 articles. In a prior issue, we have discussed the thoracic vein anatomy relevant to paroxysmal atrial fibrillation. In the present article, we focus on the atria themselves, the autonomic ganglia, and anatomic issues relevant for minimizing complications during atrial fibrillation ablation.

  7. Alternative Treatment for Bleeding Peristomal Varices: Percutaneous Parastomal Embolization

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Pabon-Ramos, Waleska M., E-mail: waly.pr@duke.edu; Niemeyer, Matthew M.; Dasika, Narasimham L., E-mail: narasimh@med.umich.edu

    2013-10-15

    Purpose: To describe how peristomal varices can be successfully embolized via a percutaneous parastomal approach. Methods: The medical records of patients who underwent this procedure between December 1, 2000, and May 31, 2008, were retrospectively reviewed. Procedural details were recorded. Median fluoroscopy time and bleeding-free interval were calculated. Results: Seven patients underwent eight parastomal embolizations. The technical success rate was 88 % (one failure). All embolizations were performed with coils combined with a sclerosant, another embolizing agent, or both. Of the seven successful parastomal embolizations, there were three cases of recurrent bleeding; the median time to rebleeding was 45 daysmore » (range 26-313 days). The remaining four patients did not develop recurrent bleeding during the follow-up period; their median bleeding-free interval was 131 days (range 40-659 days). Conclusion: This case review demonstrated that percutaneous parastomal embolization is a feasible technique to treat bleeding peristomal varices.« less

  8. Acoustic radiation force impulse imaging for real-time observation of lesion development during radiofrequency ablation procedures

    NASA Astrophysics Data System (ADS)

    Fahey, Brian J.; Trahey, Gregg E.

    2005-04-01

    When performing radiofrequency ablation (RFA) procedures, physicians currently have little or no feedback concerning the success of the treatment until follow-up assessments are made days to weeks later. To be successful, RFA must induce a thermal lesion of sufficient volume to completely destroy a target tumor or completely isolate an aberrant cardiac pathway. Although ultrasound, computed tomography (CT), and CT-based fluoroscopy have found use in guiding RFA treatments, they are deficient in giving accurate assessments of lesion size or boundaries during procedures. As induced thermal lesion size can vary considerably from patient to patient, the current lack of real-time feedback during RFA procedures is troublesome. We have developed a technique for real-time monitoring of thermal lesion size during RFA procedures utilizing acoustic radiation force impulse (ARFI) imaging. In both ex vivo and in vivo tissues, ARFI imaging provided better thermal lesion contrast and better overall appreciation for lesion size and boundaries relative to conventional sonography. The thermal safety of ARFI imaging for use at clinically realistic depths was also verified through the use of finite element method models. As ARFI imaging is implemented entirely on a diagnostic ultrasound scanner, it is a convenient, inexpensive, and promising modality for monitoring RFA procedures in vivo.

  9. [New ultrasound navigational system in extracorporeal lithotripsy: decreased fluoroscopy and radiation].

    PubMed

    Abid, N; Ravier, E; Codas, R; Crouzet, S; Martin, X

    2013-09-01

    Extracorporeal shock wave lithotripsy is the most common method of treatment for kidney stones. Both fluoroscopy and ultrasound imaging can be used to locate stones, but fluoroscopy is more frequently employed. Evaluation of a new stereotaxic navigational system: the stone was located using an ultrasound probe, and its 3D location was saved. The table automatically moved to position the stone at the focal point. A real-time follow-up was possible during treatment. Our objective was to demonstrate a decrease in the use of fluoroscopy to locate kidney stones for extracorporeal shock wave lithotripsy through the use of a 3D ultrasound stone locking system. Prospective analysis of the case records of the 20 patients preceding and the 20 patients succeeding the arrival of the ultrasound stone locking system Visio-Track (EDAP-TMS). We used a Student test to compare age, BMI, kidney stone size, number of shock waves and administered energy. Patient characteristics were comparable. The average age was 55 years old and the average kidney stone size was 10.7 mm. Radiation duration was 174.8 seconds in the group without Visio-Track versus 57.1 seconds in the group with it (P<0.0001). A similar result was observed for radiation doses: 5197.25 mGy x cm2 for the group without versus 1987.6 mGy x cm2 for the group with Visio-Track (P=0.0033). The stone locking system Visio-Track reduced fluoroscopy in our first group of patients, which decreased the patient's individual absorbed irradiation dose. Copyright © 2013 Elsevier Masson SAS. All rights reserved.

  10. [Fluoroscopy dose reduction of computed tomography guided chest interventional radiology using real-time iterative reconstruction].

    PubMed

    Hasegawa, Hiroaki; Mihara, Yoshiyuki; Ino, Kenji; Sato, Jiro

    2014-11-01

    The purpose of this study was to evaluate the radiation dose reduction to patients and radiologists in computed tomography (CT) guided examinations for the thoracic region using CT fluoroscopy. Image quality evaluation of the real-time filtered back-projection (RT-FBP) images and the real-time adaptive iterative dose reduction (RT-AIDR) images was carried out on noise and artifacts that were considered to affect the CT fluoroscopy. The image standard deviation was improved in the fluoroscopy setting with less than 30 mA on 120 kV. With regard to the evaluation of artifact visibility and the amount generated by the needle attached to the chest phantom, there was no significant difference between the RT-FBP images with 120 kV, 20 mA and the RT-AIDR images with low-dose conditions (greater than 80 kV, 30 mA and less than 120 kV, 20 mA). The results suggest that it is possible to reduce the radiation dose by approximately 34% at the maximum using RT-AIDR while maintaining image quality equivalent to the RT-FBP images with 120 V, 20 mA.

  11. [Routine fluoroscopic investigations after primary bariatric surgery].

    PubMed

    Gärtner, D; Ernst, A; Fedtke, K; Jenkner, J; Schöttler, A; Reimer, P; Blüher, M; Schön, M R

    2016-03-01

    Staple line and anastomotic leakages are life-threatening complications after bariatric surgery. Upper gastrointestinal (GI) tract X-ray examination with oral administration of a water-soluble contrast agent can be used to detect leaks. The aim of this study was to evaluate the impact of routine upper GI tract fluoroscopy after primary bariatric surgery. Between January 2009 and December 2014 a total of 658 bariatric interventions were carried out of which 442 were primary bariatric operations. Included in this single center study were 307 sleeve gastrectomies and 135 Roux-en-Y gastric bypasses. Up to December 2012 upper GI tract fluoroscopy was performed routinely between the first and third postoperative days and the detection of leakages was evaluated. In the investigation period 8 leakages (2.6 %) after sleeve gastrectomy, 1 anastomotic leakage in gastrojejunostomy and 1 in jejunojejunostomy after Roux-en-Y gastric bypass occurred. All patients developed clinical symptoms, such as abdominal pain, tachycardia or fever. In one case the leakage was detected by upper GI fluoroscopy and in nine cases radiological findings were unremarkable. No leakages were detected in asymptomatic patients. Routine upper GI fluoroscopy is not recommended for uneventful postoperative courses after primary bariatric surgery.

  12. [Evaluation of Radiation Dose during Stent-graft Treatment Using a Hybrid Operating Room System].

    PubMed

    Haga, Yoshihiro; Chida, Kouichi; Kaga, Yuji; Saitou, Kazuhisa; Arai, Takeshi; Suzuki, Shinichi; Iwaya, Yoshimi; Kumasaka, Eriko; Kataoka, Nozomi; Satou, Naoto; Abe, Mitsuya

    2015-12-01

    In recent years, aortic aneurysm treatment with stent graft grafting in the X-ray fluoroscopy is increasing. This is an endovascular therapy, because it is a treatment which includes the risk of radiation damage, having to deal with radiation damage, to know in advance is important. In this study, in order to grasp the trend of exposure stent graft implantation in a hybrid operating room (OR) system, focusing on clinical data (entrance skin dose and fluoroscopy time), was to count the total. In TEVAR and EVAR, fluoroscopy time became 13.40 ± 7.27 minutes, 23.67 ± 11.76 minutes, ESD became 0.87 ± 0.41 mGy, 1.11 ± 0.57 mGy. (fluoroscopy time of EVAR was 2.0 times than TEVAR. DAP of EVAR was 1.2 times than TEVAR.) When using the device, adapted lesions and usage are different. This means that care changes in exposure-related factors. In this study, exposure trends of the stent graft implantation was able to grasp. It can be a helpful way to reduce/optimize the radiation dose in a hybrid OR system.

  13. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ogawa, Yukihisa, E-mail: yukky.oct.22@gmail.com; Hamaguchi, Shingo; Nishimaki, Hiroshi

    BackgroundEndovascular aortic repair (EVAR) requires further intervention in 20-30 % of cases, often due to type II endoleak (T2EL). Management options for T2EL include transarterial embolization, direct puncture (DP), or transcaval embolization. We report the case of an 80-year-old man with T2EL who successfully underwent DP embolization.MethodsEmbolization by DP was performed with a transpedicular approach using an isocenter puncture (ISOP) method. An isocenter marker (ICM) was placed at a site corresponding to the aneurysm sac on fluoroscopy in two directions (frontal and lateral views). A vertebroplasty needle was inserted tangentially to the ICM under fluoroscopy and advanced to the anterior wallmore » of the vertebral body. A 20 cm-length, 20-G-PTCD needle was inserted through the outer needle of the 13-G needle and advanced to the ICM. Sac embolization using 25 % N-buty-2-cyanoacrylate diluted with Lipiodol was performed. After complete embolization, rotational DA confirmed good filling of the sac with Lipiodol. The outer cannula and 13-G needle were removed and the procedure was completed.ResultsThe patient was discharged the next day. Contrast-enhanced computed tomography 1 and 8 months later showed no Lipiodol washout in the aneurysm sac, no endoleak recurrence, and no expansion of the excluded aneurysm.ConclusionDP with a transpedicular approach using ISOP may be useful when translumbar and transabdominal approaches prove difficult.« less

  14. Dynamic flat panel detector versus image intensifier in cardiac imaging: dose and image quality

    NASA Astrophysics Data System (ADS)

    Vano, E.; Geiger, B.; Schreiner, A.; Back, C.; Beissel, J.

    2005-12-01

    The practical aspects of the dosimetric and imaging performance of a digital x-ray system for cardiology procedures were evaluated. The system was configured with an image intensifier (II) and later upgraded to a dynamic flat panel detector (FD). Entrance surface air kerma (ESAK) to phantoms of 16, 20, 24 and 28 cm of polymethyl methacrylate (PMMA) and the image quality of a test object were measured. Images were evaluated directly on the monitor and with numerical methods (noise and signal-to-noise ratio). Information contained in the DICOM header for dosimetry audit purposes was also tested. ESAK values per frame (or kerma rate) for the most commonly used cine and fluoroscopy modes for different PMMA thicknesses and for field sizes of 17 and 23 cm for II, and 20 and 25 cm for FD, produced similar results in the evaluated system with both technologies, ranging between 19 and 589 µGy/frame (cine) and 5 and 95 mGy min-1 (fluoroscopy). Image quality for these dose settings was better for the FD version. The 'study dosimetric report' is comprehensive, and its numerical content is sufficiently accurate. There is potential in the future to set those systems with dynamic FD to lower doses than are possible in the current II versions, especially for digital cine runs, or to benefit from improved image quality.

  15. Percutaneous CT and Fluoroscopy-Guided Screw Fixation of Pathological Fractures in the Shoulder Girdle: Technical Report of 3 Cases.

    PubMed

    Garnon, Julien; Koch, Guillaume; Ramamurthy, Nitin; Caudrelier, Jean; Rao, Pramod; Tsoumakidou, Georgia; Cazzato, Roberto Luigi; Gangi, Afshin

    2016-09-01

    To review our initial experience with percutaneous CT and fluoroscopy-guided screw fixation of pathological shoulder-girdle fractures. Between May 2014 and June 2015, three consecutive oncologic patients (mean age 65 years; range 57-75 years) with symptomatic pathological shoulder-girdle fractures unsuitable for surgery and radiotherapy underwent percutaneous image-guided screw fixation. Fractures occurred through metastases (n = 2) or a post-ablation cavity (n = 1). Mechanical properties of osteosynthesis were adjudged superior to stand-alone cementoplasty in each case. Cannulated screws were placed under combined CT and fluoroscopic guidance with complementary radiofrequency ablation or cementoplasty to optimise local palliation and secure screw fixation, respectively, in two cases. Follow-up was undertaken every few weeks until mortality or most recent appointment. Four pathological fractures were treated in three patients (2 acromion, 1 clavicular, 1 coracoid). Mean size of associated lesion was 2.6 cm (range 1-4.5 cm). Technical success was achieved in all cases (100 %), without complications. Good palliation and restoration of mobility were observed in two cases at 2-3 months; one case could not be followed due to early post-procedural oncologic mortality. Percutaneous image-guided shoulder-girdle osteosynthesis appears technically feasible with good short-term efficacy in this complex patient subset. Further studies are warranted to confirm these promising initial results.

  16. Study of scattered radiation during fluoroscopy in hip surgery*

    PubMed Central

    Lesyuk, Oksana; Sousa, Patrick Emmanuel; Rodrigues, Sónia Isabel do Espirito Santo; Abrantes, António Fernando; de Almeida, Rui Pedro Pereira; Pinheiro, João Pedro; Azevedo, Kevin Barros; Ribeiro, Luís Pedro Vieira

    2016-01-01

    Objective To measure the scattered radiation dose at different positions simulating hip surgery. Materials and Methods We simulated fluoroscopy-assisted hip surgery in order to study the distribution of scattered radiation in the operating room. To simulate the patient, we used a anthropomorphic whole-body phantom, and we used an X-ray-specific detector to quantify the radiation. Radiographs were obtained with a mobile C-arm X-ray system in continuous scan mode, with the tube at 0º (configuration 1) or 90º (configuration 2). The operating parameters employed (voltage, current, and exposure time) were determined by a statistical analysis based on the observation of orthopedic surgical procedures involving the hip. Results For all measurements, higher exposures were observed in configuration 2. In the measurements obtained as a function of height, the maximum dose rates observed were 1.167 (± 0.023) µSv/s and 2.278 (± 0.023) µSv/s in configurations 1 and 2, respectively, corresponding to the chest level of health care professionals within the operating room. Proximal to the patient, the maximum values were recorded in the position occupied by the surgeon. Conclusion We can conclude that, in the scenario under study, health care professionals workers are exposed to low levels of radiation, and that those levels can be reduced through the use of personal protective equipment. PMID:27777477

  17. Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lin, Pei-Jan P., E-mail: Pei-Jan.Lin@vcuhealth.org; Schueler, Beth A.; Balter, Stephen

    2015-12-15

    Due to the proliferation of disciplines employing fluoroscopy as their primary imaging tool and the prolonged extensive use of fluoroscopy in interventional and cardiovascular angiography procedures, “dose-area-product” (DAP) meters were installed to monitor and record the radiation dose delivered to patients. In some cases, the radiation dose or the output value is calculated, rather than measured, using the pertinent radiological parameters and geometrical information. The AAPM Task Group 190 (TG-190) was established to evaluate the accuracy of the DAP meter in 2008. Since then, the term “DAP-meter” has been revised to air kerma-area product (KAP) meter. The charge of TGmore » 190 (Accuracy and Calibration of Integrated Radiation Output Indicators in Diagnostic Radiology) has also been realigned to investigate the “Accuracy and Calibration of Integrated Radiation Output Indicators” which is reflected in the title of the task group, to include situations where the KAP may be acquired with or without the presence of a physical “meter.” To accomplish this goal, validation test protocols were developed to compare the displayed radiation output value to an external measurement. These test protocols were applied to a number of clinical systems to collect information on the accuracy of dose display values in the field.« less

  18. Magnetically Assisted Remote-controlled Endovascular Catheter for Interventional MR Imaging: In Vitro Navigation at 1.5 T versus X-ray Fluoroscopy

    PubMed Central

    Losey, Aaron D.; Lillaney, Prasheel; Martin, Alastair J.; Cooke, Daniel L.; Wilson, Mark W.; Thorne, Bradford R. H.; Sincic, Ryan S.; Arenson, Ronald L.; Saeed, Maythem

    2014-01-01

    Purpose To compare in vitro navigation of a magnetically assisted remote-controlled (MARC) catheter under real-time magnetic resonance (MR) imaging with manual navigation under MR imaging and standard x-ray guidance in endovascular catheterization procedures in an abdominal aortic phantom. Materials and Methods The 2-mm-diameter custom clinical-grade microcatheter prototype with a solenoid coil at the distal tip was deflected with a foot pedal actuator used to deliver 300 mA of positive or negative current. Investigators navigated the catheter into branch vessels in a custom cryogel abdominal aortic phantom. This was repeated under MR imaging guidance without magnetic assistance and under conventional x-ray fluoroscopy. MR experiments were performed at 1.5 T by using a balanced steady-state free precession sequence. The mean procedure times and percentage success data were determined and analyzed with a linear mixed-effects regression analysis. Results The catheter was clearly visible under real-time MR imaging. One hundred ninety-two (80%) of 240 turns were successfully completed with magnetically assisted guidance versus 144 (60%) of 240 turns with nonassisted guidance (P < .001) and 119 (74%) of 160 turns with standard x-ray guidance (P = .028). Overall mean procedure time was shorter with magnetically assisted than with nonassisted guidance under MR imaging (37 seconds ± 6 [standard error of the mean] vs 55 seconds ± 3, P < .001), and time was comparable between magnetically assisted and standard x-ray guidance (37 seconds ± 6 vs 44 seconds ± 3, P = .045). When stratified by angle of branch vessel, magnetic assistance was faster than nonassisted MR guidance at turns of 45°, 60°, and 75°. Conclusion In this study, a MARC catheter for endovascular navigation under real-time MR imaging guidance was developed and tested. For catheterization of branch vessels arising at large angles, magnetically assisted catheterization was faster than manual catheterization under MR imaging guidance and was comparable to standard x-ray guidance. © RSNA, 2014 Online supplemental material is available for this article. PMID:24533872

  19. Antegrade Ureteral Stenting is a Good Alternative for the Retrograde Approach.

    PubMed

    van der Meer, Rutger W; Weltings, Saskia; van Erkel, Arian R; Roshani, Hossain; Elzevier, Henk W; van Dijk, Lukas C; van Overhagen, Hans

    2017-07-01

    Double J (JJ) stents for treating obstructive ureteral pathology are generally inserted through a retrograde route with cystoscopic guidance. Antegrade percutaneous insertion using fluoroscopy can be performed alternatively but is less known. Indications, success rate and complications of antegrade ureteral stenting were evaluated. Data of consecutive patients in which antegrade ureteral stenting was performed were retrospectively analysed using the radiology information system and patient records. Patient characteristics, details of the antegrade JJ stent insertion procedure and registered complications were collected. Furthermore, it was investigated if prior to the antegrade procedure a retrograde attempt for JJ stent insertion was performed. Total 130 attempts for antegrade JJ stent insertion were performed in 100 patients. A percutaneous nephrostomy catheter had already been placed in the majority of kidneys (n = 109) for initial treatment of hydronephrosis. Most prevelant indication for a JJ stent was obstructive ureteral pathology due to malignancy (n = 63). A JJ stent was successfully inserted in 125 of 130 procedures. In 21 cases, previous retrograde ureteral stenting had failed but, subsequent antegrade ureteral stenting was successful. There were 8 procedure related complications; 6 infections, 1 false tract and 1 malposition. Antegrade percutaneous insertion of a JJ stent is a good alternative for retrograde insertion.

  20. How well can step-off and gap distances be reduced when treating intra-articular distal radius fractures with fragment specific fixation when using fluoroscopy.

    PubMed

    Thiart, M; Ikram, A; Lamberts, R P

    2016-12-01

    Although fragment specific fixation has proved to be an effective treatment regime, it has not been established how successfully this treatment could be performed using fluoroscopy and what the added value of arthroscopy could be. Establish gap and step-off distances after in intra-articular distal radius fractures that have been treated with fragment specific fixation while using fluoroscopy. Forty-four patients with an intra-articular distal radius fracture were treated with fragment specific fixation while using fluoroscopy. After the treatment of the intra-articular distal radius fracture with fragment specific fixation and the use of fluoroscopy, but before the completion of the surgical intervention, all gap, and step-off distances were determined by using arthroscopy. In addition, the joint was checked for any other wrist pathologies. Arthroscopy after the surgical intervention showed that in 37 patients no gap distances could be detected, while in six patients a gap distance of≤2mm was found and in one patient, a gap distance of 3mm. Similarly, arthroscopy revealed no step-off distances in 33 patients, while in 11 patients a step-off distance of≤2mm was found. Although additional wrist pathologies were found in 48% of our population, only one patient needed surgical intervention. Three months after the surgical intervention wrist flexion was 41±10°, wrist extension 51±17°, ulnar deviation 19±10°, radial deviation 32±12° while patients could pronate and supinate their wrist to 85±5° and 74±20°, respectively. Intra-articular distal radius fractures can be treated successfully with fragment specific fixation and the use of fluoroscopy. As almost all gap and step-off distances could be reduced to an acceptable level, the scope for arthroscopy to further improve this treatment regime is limited. The functional outcome scores that were found 3 months after the surgical intervention were similar to what has been reported in other studies using different treatment option. These findings suggest that fragment specific fixation is a good alternative for treating intra-articular distal radius fractures. As in most cases, only fluoroscopy is needed for fragment specific fixation, this treatment technique is a good treatment option for resource-limited hospitals, setting who do not have access to arthroscopy. III, case-control study. Copyright © 2016 Elsevier Masson SAS. All rights reserved.

  1. Learning, Realizability and Games in Classical Arithmetic

    NASA Astrophysics Data System (ADS)

    Aschieri, Federico

    2010-12-01

    In this dissertation we provide mathematical evidence that the concept of learning can be used to give a new and intuitive computational semantics of classical proofs in various fragments of Predicative Arithmetic. First, we extend Kreisel modified realizability to a classical fragment of first order Arithmetic, Heyting Arithmetic plus EM1 (Excluded middle axiom restricted to Sigma^0_1 formulas). We introduce a new realizability semantics we call "Interactive Learning-Based Realizability". Our realizers are self-correcting programs, which learn from their errors and evolve through time. Secondly, we extend the class of learning based realizers to a classical version PCFclass of PCF and, then, compare the resulting notion of realizability with Coquand game semantics and prove a full soundness and completeness result. In particular, we show there is a one-to-one correspondence between realizers and recursive winning strategies in the 1-Backtracking version of Tarski games. Third, we provide a complete and fully detailed constructive analysis of learning as it arises in learning based realizability for HA+EM1, Avigad's update procedures and epsilon substitution method for Peano Arithmetic PA. We present new constructive techniques to bound the length of learning processes and we apply them to reprove - by means of our theory - the classic result of Godel that provably total functions of PA can be represented in Godel's system T. Last, we give an axiomatization of the kind of learning that is needed to computationally interpret Predicative classical second order Arithmetic. Our work is an extension of Avigad's and generalizes the concept of update procedure to the transfinite case. Transfinite update procedures have to learn values of transfinite sequences of non computable functions in order to extract witnesses from classical proofs.

  2. Automatic Localization of Vertebral Levels in X-Ray Fluoroscopy Using 3D-2D Registration: A Tool to Reduce Wrong-Site Surgery

    PubMed Central

    Otake, Y.; Schafer, S.; Stayman, J. W.; Zbijewski, W.; Kleinszig, G.; Graumann, R.; Khanna, A. J.; Siewerdsen, J. H.

    2012-01-01

    Surgical targeting of the incorrect vertebral level (“wrong-level” surgery) is among the more common wrong-site surgical errors, attributed primarily to a lack of uniquely identifiable radiographic landmarks in the mid-thoracic spine. Conventional localization method involves manual counting of vertebral bodies under fluoroscopy, is prone to human error, and carries additional time and dose. We propose an image registration and visualization system (referred to as LevelCheck), for decision support in spine surgery by automatically labeling vertebral levels in fluoroscopy using a GPU-accelerated, intensity-based 3D-2D (viz., CT-to-fluoroscopy) registration. A gradient information (GI) similarity metric and CMA-ES optimizer were chosen due to their robustness and inherent suitability for parallelization. Simulation studies involved 10 patient CT datasets from which 50,000 simulated fluoroscopic images were generated from C-arm poses selected to approximate C-arm operator and positioning variability. Physical experiments used an anthropomorphic chest phantom imaged under real fluoroscopy. The registration accuracy was evaluated as the mean projection distance (mPD) between the estimated and true center of vertebral levels. Trials were defined as successful if the estimated position was within the projection of the vertebral body (viz., mPD < 5mm). Simulation studies showed a success rate of 99.998% (1 failure in 50,000 trials) and computation time of 4.7 sec on a midrange GPU. Analysis of failure modes identified cases of false local optima in the search space arising from longitudinal periodicity in vertebral structures. Physical experiments demonstrated robustness of the algorithm against quantum noise and x-ray scatter. The ability to automatically localize target anatomy in fluoroscopy in near-real-time could be valuable in reducing the occurrence of wrong-site surgery while helping to reduce radiation exposure. The method is applicable beyond the specific case of vertebral labeling, since any structure defined in pre-operative (or intra-operative) CT or cone-beam CT can be automatically registered to the fluoroscopic scene. PMID:22864366

  3. Detecting condylar contact loss using single-plane fluoroscopy: a comparison with in vivo force data and in vitro bi-plane data.

    PubMed

    Prins, A H; Kaptein, B L; Banks, S A; Stoel, B C; Nelissen, R G H H; Valstar, E R

    2014-05-07

    Knee contact mechanics play an important role in knee implant failure and wear mechanics. Femoral condylar contact loss in total knee arthroplasty has been reported in some studies and it is considered to potentially induce excessive wear of the polyethylene insert.Measuring in vivo forces applied to the tibial plateau with an instrumented prosthesis is a possible approach to assess contact loss in vivo, but this approach is not very practical. Alternatively, single-plane fluoroscopy and pose estimation can be used to derive the relative pose of the femoral component with respect to the tibial plateau and estimate the distance from the medial and lateral parts of the femoral component towards the insert. Two measures are reported in the literature: lift-off is commonly defined as the difference in distance between the medial and lateral condyles of the femoral component with respect to the tibial plateau; separation is determined by the closest distance of each condyle towards the polyethylene insert instead of the tibia plateau.In this validation study, lift-off and separation as measured with single-plane fluoroscopy are compared to in vivo contact forces measured with an instrumented knee implant. In a phantom study, lift-off and separation were compared to measurements with a high quality bi-plane measurement.The results of the in vivo contact-force experiment demonstrate a large discrepancy between single-plane fluoroscopy and the in vivo force data: single-plane fluoroscopy measured up to 5.1mm of lift-off or separation, whereas the force data never showed actual loss of contact. The phantom study demonstrated that the single-plane setup could introduce an overestimation of 0.22mm±±0.36mm. Correcting the out-of-plane position resulted in an underestimation of medial separation by -0.20mm±±0.29mm.In conclusion, there is a discrepancy between the in vivo force data and single-plane fluoroscopic measurements. Therefore contact loss may not always be determined reliably by single plane fluoroscopy analysis. Copyright © 2014 Elsevier Ltd. All rights reserved.

  4. SU-G-JeP4-12: Real-Time Organ Motion Monitoring Using Ultrasound and KV Fluoroscopy During Lung SBRT Delivery

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Omari, E; Tai, A; Li, X

    Purpose: Real-time ultrasound monitoring during SBRT is advantageous in understanding and identifying motion irregularities which may cause geometric misses. In this work, we propose to utilize real-time ultrasound to track the diaphragm in conjunction with periodical kV fluoroscopy to monitor motion of tumor or landmarks during SBRT delivery. Methods: Transabdominal Ultrasound (TAUS) b-mode images were collected from 10 healthy volunteers using the Clarity Autoscan System (Elekta). The autoscan transducer, which has a center frequency of 5 MHz, was utilized for the scans. The acquired images were contoured using the Clarity Automatic Fusion and Contouring workstation software. Monitoring sessions of 5more » minute length were observed and recorded. The position correlation between tumor and diaphragm could be established with periodic kV fluoroscopy periodically acquired during treatment with Elekta XVI. We acquired data using a tissue mimicking ultrasound phantom with embedded spheres placed on a motion stand using ultrasound and kV Fluoroscopy. MIM software was utilized for image fusion. Correlation of diaphragm and target motion was also validated using 4D-MRI and 4D-CBCT. Results: The diaphragm was visualized as a hyperechoic region on the TAUS b-mode images. Volunteer set-up can be adjusted such that TAUS probe will not interfere with treatment beams. A segment of the diaphragm was contoured and selected as our tracking structure. Successful monitoring sessions of the diaphragm were recorded. For some volunteers, diaphragm motion over 2 times larger than the initial motion has been observed during tracking. For the phantom study, we were able to register the 2D kV Fluoroscopy with the US images for position comparison. Conclusion: We demonstrated the feasibility of tracking the diaphragm using real-time ultrasound. Real-time tracking can help in identifying such irregularities in the respiratory motion which is correlated to tumor motion. We also showed the feasibility of acquiring 2D KV Fluoroscopy and registering the images with Ultrasound.« less

  5. Ultrasound-guided drainages and sclerotherapy.

    PubMed

    Lohela, P

    2002-02-01

    Ultrasonically guided fluid collection and abscess drainage have become routine procedures in various parts of the body. In most cases ultrasound is the only imaging and guidance modality needed; however, it is of the utmost importance to remember that CT and fluoroscopy with contrast often give invaluable information when the true extent of the process has to be determined and when assessing the safest route for the catheter in anatomically complicated areas. The importance of irrigation of the abscess cavity with fluids and the ready use of urokinase should be emphasized. Ethanol sclerotherapy is a simple and safe procedure to treat symptomatic hepatic or renal cysts. Parathyroid adenomas and cysts, as well as thyroid cysts, can also be treated with ethanol sclerotherapy in selected cases. Purified mineral talc has been used in pleurodesis and hydrocele sclerotherapy, whereas doxycycline or ethanol is used for postoperative lymphoceles. Both abscess drainages and sclerotherapy procedures are minimally invasive, simple, safe, inexpensive and reasonably efficacious treatment in many clinical instances and may be at least an alternative to surgical treatment, often offering significant advantages over surgery.

  6. A method to reduce patient's eye lens dose in neuro-interventional radiology procedures

    NASA Astrophysics Data System (ADS)

    Safari, M. J.; Wong, J. H. D.; Kadir, K. A. A.; Sani, F. M.; Ng, K. H.

    2016-08-01

    Complex and prolonged neuro-interventional radiology procedures using the biplane angiography system increase the patient's risk of radiation-induced cataract. Physical collimation is the most effective way of reducing the radiation dose to the patient's eye lens, but in instances where collimation is not possible, an attenuator may be useful in protecting the eyes. In this study, an eye lens protector was designed and fabricated to reduce the radiation dose to the patients' eye lens during neuro-interventional procedures. The eye protector was characterised before being tested on its effectiveness in a simulated aneurysm procedure on an anthropomorphic phantom. Effects on the automatic dose rate control (ADRC) and image quality are also evaluated. The eye protector reduced the radiation dose by up to 62.1% at the eye lens. The eye protector is faintly visible in the fluoroscopy images and increased the tube current by a maximum of 3.7%. It is completely invisible in the acquisition mode and does not interfere with the clinical procedure. The eye protector placed within the radiation field of view was able to reduce the radiation dose to the eye lens by direct radiation beam of the lateral x-ray tube with minimal effect on the ADRC system.

  7. Editor's choice--Use of disposable radiation-absorbing surgical drapes results in significant dose reduction during EVAR procedures.

    PubMed

    Kloeze, C; Klompenhouwer, E G; Brands, P J M; van Sambeek, M R H M; Cuypers, P W M; Teijink, J A W

    2014-03-01

    Because of the increasing number of interventional endovascular procedures with fluoroscopy and the corresponding high annual dose for interventionalists, additional dose-protecting measures are desirable. The purpose of this study was to evaluate the effect of disposable radiation-absorbing surgical drapes in reducing scatter radiation exposure for interventionalists and supporting staff during an endovascular aneurysm repair (EVAR) procedure. This was a randomized control trial in which 36 EVAR procedures were randomized between execution with and without disposable radiation-absorbing surgical drapes (Radpad: Worldwide Innovations & Technologies, Inc., Kansas City, US, type 5511A). Dosimetric measurements were performed on the interventionalist (hand and chest) and theatre nurse (chest) with and without the use of the drapes to obtain the dose reduction and effect on the annual dose caused by the drapes. Use of disposable radiation-absorbing surgical drapes resulted in dose reductions of 49%, 55%, and 48%, respectively, measured on the hand and chest of the interventionalist and the chest of the theatre nurse. The use of disposable radiation-absorbing surgical drapes significantly reduces scatter radiation exposure for both the interventionalist and the supporting staff during EVAR procedures. Copyright © 2013 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

  8. Percutaneous magnetic resonance imaging-guided bone tumor management and magnetic resonance imaging-guided bone therapy.

    PubMed

    Sequeiros, Roberto Blanco; Fritz, Jan; Ojala, Risto; Carrino, John A

    2011-08-01

    Magnetic resonance imaging (MRI) is promising tool for image-guided therapy. In musculoskeletal setting, image-guided therapy is used to direct diagnostic and therapeutic procedures and to steer patient management. Studies have demonstrated that MRI-guided interventions involving bone, soft tissue, joints, and intervertebral disks are safe and in selected indications can be the preferred action to manage clinical situation. Often, these procedures are technically similar to those performed in other modalities (computed tomography, fluoroscopy) for bone and soft tissue lesions. However, the procedural perception to the operator can be very different to other modalities because of the vastly increased data.Magnetic resonance imaging guidance is particularly advantageous should the lesion not be visible by other modalities, for selective lesion targeting, intra-articular locations, cyst aspiration, and locations adjacent to surgical hardware. Palliative tumor-related pain management such as ablation therapy forms a subset of procedures that are frequently performed under MRI. Another suitable entity for MRI guidance are the therapeutic percutaneous osseous or joint-related benign or reactive conditions such as osteoid osteoma, epiphyseal bone bridging, osteochondritis dissecans, bone cysts, localized bone necrosis, and posttraumatic lesions. In this article, we will describe in detail the technical aspects of performing MRI-guided therapeutic musculoskeletal procedures as well as the clinical indications.

  9. Descriptive Cadaveric Study Comparing the Accuracy of Ultrasound Versus Fluoroscopic Guidance for First Sacral Transforaminal Injections: A Comparison Study.

    PubMed

    Thompson, Bradley F; Pingree, Matthew J; Qu, Wenchun; Murthy, Naveen S; Lachman, Nirusha; Hurdle, Mark Friedrich

    2018-04-01

    Ultrasound is rarely used for guiding lumbosacral epidural steroid injections due to its technical limitations. For example, sonographic imaging lacks the ability to confirm epidural spread and identify vascular uptake. The perceived risk that these limitations pose to human subjects has precluded any large scale clinical trials to date. To compare the accuracy of ultrasound versus fluoroscopic guidance for first sacral transforaminal epidural injections. Cadaveric comparative study using dichotomous outcomes. A fluoroscopy suite and anatomic laboratory at an academic medical center. Four unembalmed adult human cadavers with no history of spinal surgery. Eight sites were injected twice by one interventionalist, using fluoroscopic and ultrasound guidance. In the fluoroscopy arm, contrast spread was assessed using computed tomography. In the ultrasound arm, latex spread was assessed using gross anatomic dissection. Any visible evidence of epidural spread constituted a positive result. Comparison of the success of obtaining epidural contrast flow was the primary outcome measure. Secondary outcome measures included average duration, rate of intravascular uptake, and quantity of intravascular uptake. All injections performed in both the ultrasound arm and the fluoroscopy arm had positive epidural spread. The average duration was 3.03 minutes with fluoroscopy and 4.76 minutes with ultrasound. The rate of intravascular uptake was 37.5% with fluoroscopy and 50% with ultrasound. Within the ultrasound arm, greater intravascular spread and duration variability were recorded. Although ultrasonography can provide reliable image guidance for cannulating the first sacral foramen in cadavers, it would have limited clinical utility due to its inability to visualize relevant neurovascular structures deep to the osseus roof and exclude intravascular uptake. IV. Copyright © 2018 American Academy of Physical Medicine and Rehabilitation. Published by Elsevier Inc. All rights reserved.

  10. Optimal angle of needle insertion for fluoroscopy-guided transforaminal epidural injection of L5.

    PubMed

    Ra, In-Hoo; Min, Woo-Kie

    2015-06-01

    Unlike other sites, there is difficulty in performing TFESI at the L5-S1 level because the iliac crest is an obstacle to needle placement. The objective of this study was to identify the optimal angle of fluoroscopy for insertion and advancement of a needle during L5 TEFSI. We conducted an observational study of patients undergoing fluoroscopy-guided L5 TFESI in the prone position. A total of 80 patients (40 men and 40 women) with radiating pain of lower limbs were enrolled. During TFESI, we measured the angle at which the L5 vertebral body forms a rectangular shape and compared men and women. Then, we measured area of safe triangle in tilting angle of fluoroscopy from 15° to 35° and compared men and women. The mean cephalocaudal angle, where the vertebral body takes the shape of a rectangle, was 11.0° in men and 13.9° in women (P = 0.007). In men, the triangular area was maximal at 18.3 mm² with an oblique view angle of 25°. In women, the area was maximal at 23.6 mm² with an oblique view angle of 30°. At an oblique view angle of 30° and 35°, the area was significantly greater in women (P < 0.05). When TFESI is performed at the L5 region in the prone position, placement of fluoroscopy at a cephalocaudal angle of 11.0° and an oblique angle of 25° in men and cephalocaudal angle of 13.9° and an oblique angle of 30° in women would be most reasonable. © 2014 World Institute of Pain.

  11. Double needle technique: an alternative method for performing difficult sacroiliac joint injections.

    PubMed

    Gupta, Sanjeeva

    2011-01-01

    The sacroiliac joint (SIJ) is a common source of low back pain. The most appropriate method of confirming SIJ pain is to inject local anesthesia into the joint to find out if the pain decreases. Unfortunately, although the SIJ is a large joint, it can be difficult to enter due to the complex nature of the joint and variations in anatomy. In my experience a double needle technique for sacroiliac joint injection can increase the chances of accurate injection into the SIJ in difficult cases. After obtaining appropriate fluoroscopic images, the tip of the needle is advanced into the SIJ. Once the tip of the needle is correctly placed, its position is checked under continuous fluoroscopy while moving the C-arm in the right and left oblique directions (dynamic fluoroscopy). On dynamic fluoroscopy the tip of the needle should remain within the joint line and not appear to be on the bone. If the tip of the needle appears to be on the bone a new joint line will need to be identified (the most translucent area through the joint) by dynamic fluoroscopy and another needle advanced into the newly identified joint line. Dynamic fluoroscopy is repeated again to confirm that the tip of the second needle remains within the joint line. Once both needles are in place contrast dye is injected through the needle that is most likely to be in the SIJ. If the contrast dye spread is not satisfactory then it is injected through the other needle. I have used this technique in 10 patients and found it very helpful in accurately performing SIJ injection which can at times be challenging.

  12. An Outcome and Cost Analysis Comparing Single-Level Minimally Invasive Transforaminal Lumbar Interbody Fusion Using Intraoperative Fluoroscopy versus Computed Tomography-Guided Navigation.

    PubMed

    Khanna, Ryan; McDevitt, Joseph L; Abecassis, Zachary A; Smith, Zachary A; Koski, Tyler R; Fessler, Richard G; Dahdaleh, Nader S

    2016-10-01

    Minimally invasive transforaminal lumbar interbody fusion (TLIF) has undergone significant evolution since its conception as a fusion technique to treat lumbar spondylosis. Minimally invasive TLIF is commonly performed using intraoperative two-dimensional fluoroscopic x-rays. However, intraoperative computed tomography (CT)-based navigation during minimally invasive TLIF is gaining popularity for improvements in visualizing anatomy and reducing intraoperative radiation to surgeons and operating room staff. This is the first study to compare clinical outcomes and cost between these 2 imaging techniques during minimally invasive TILF. For comparison, 28 patients who underwent single-level minimally invasive TLIF using fluoroscopy were matched to 28 patients undergoing single-level minimally invasive TLIF using CT navigation based on race, sex, age, smoking status, payer type, and medical comorbidities (Charlson Comorbidity Index). The minimum follow-up time was 6 months. The 2 groups were compared in regard to clinical outcomes and hospital reimbursement from the payer perspective. Average surgery time, anesthesia time, and hospital length of stay were similar for both groups, but average estimated blood loss was lower in the fluoroscopy group compared with the CT navigation group (154 mL vs. 262 mL; P = 0.016). Oswestry Disability Index, back visual analog scale, and leg visual analog scale scores similarly improved in both groups (P > 0.05) at 6-month follow-up. Cost analysis showed that average hospital payments were similar in the fluoroscopy versus the CT navigation groups ($32,347 vs. $32,656; P = 0.925) as well as payments for the operating room (P = 0.868). Single minimally invasive TLIF performed with fluoroscopy versus CT navigation showed similar clinical outcomes and cost at 6 months. Copyright © 2016 Elsevier Inc. All rights reserved.

  13. Cone-beam computed tomography fusion and navigation for real-time positron emission tomography-guided biopsies and ablations: a feasibility study.

    PubMed

    Abi-Jaoudeh, Nadine; Mielekamp, Peter; Noordhoek, Niels; Venkatesan, Aradhana M; Millo, Corina; Radaelli, Alessandro; Carelsen, Bart; Wood, Bradford J

    2012-06-01

    To describe a novel technique for multimodality positron emission tomography (PET) fusion-guided interventions that combines cone-beam computed tomography (CT) with PET/CT before the procedure. Subjects were selected among patients scheduled for a biopsy or ablation procedure. The lesions were not visible with conventional imaging methods or did not have uniform uptake on PET. Clinical success was defined by adequate histopathologic specimens for molecular profiling or diagnosis and by lack of enhancement on follow-up imaging for ablation procedures. Time to target (time elapsed between the completion of the initial cone-beam CT scan and first tissue sample or treatment), total procedure time (time from the moment the patient was on the table until the patient was off the table), and number of times the needle was repositioned were recorded. Seven patients underwent eight procedures (two ablations and six biopsies). Registration and procedures were completed successfully in all cases. Clinical success was achieved in all biopsy procedures and in one of the two ablation procedures. The needle was repositioned once in one biopsy procedure only. On average, the time to target was 38 minutes (range 13-54 min). Total procedure time was 95 minutes (range 51-240 min, which includes composite ablation). On average, fluoroscopy time was 2.5 minutes (range 1.3-6.2 min). An integrated cone-beam CT software platform can enable PET-guided biopsies and ablation procedures without the need for additional specialized hardware. Copyright © 2012 SIR. Published by Elsevier Inc. All rights reserved.

  14. CT Guided Bone Biopsy Using a Battery Powered Intraosseous Device

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Schnapauff, Dirk, E-mail: dirk.schnapauff@charite.de; Marnitz, Tim, E-mail: tim.marnitz@charite.de; Freyhardt, Patrick, E-mail: Patrick.freyhardt@charite.de

    2013-10-15

    Purpose: To evaluate the feasibility of a battery powered intraosseous device to perform CT-fluoroscopy guided bone biopsy. Methods: Retrospective analysis of 12 patients in whom bone specimen were acquired from different locations under CT-fluoroscopy guidance using the OnControl bone marrow biopsy system (OBM, Vidacare, Shavano Park, TX, USA). Data of the 12 were compared to a historic cohort in whom the specimen were acquired using the classic Jamshidi Needle, as reference needle using manual force for biopsy. Results: Technical success was reached in 11 of 12 cases, indicated by central localisation of the needle within the target lesion. All specimenmore » sampled were sufficient for histopathological workup. Compared to the historical cohort the time needed for biopsy decreased significantly from 13 {+-} 6 to 6 {+-} 4 min (P = 0.0001). Due to the shortened intervention time the radiation dose (CTDI) during CT-fluoroscopy was lowered significantly from 169 {+-} 87 to 111 {+-} 54 mGy Multiplication-Sign cm (P = 0.0001). Interventional radiologists were confident with the performance of the needle especially when using in sclerotic or osteoblastic lesions. Conclusion: The OBM is an attractive support for CT-fluoroscopy guided bone biopsy which is safe tool and compared to the classical approach using the Jamshidi needle leading to significantly reduced intervention time and radiation exposure.« less

  15. Initial experience of a novel mapping system combined with remote magnetic navigation in the catheter ablation of atrial fibrillation.

    PubMed

    Lin, Changjian; Pehrson, Steen; Jacobsen, Peter Karl; Chen, Xu

    2017-12-01

    There have been advancements of sophisticated mapping systems used for ablation procedures over the last decade. Utilization of these novel mapping systems in combination with remote magnetic navigation (RMN) needs to be established. We investigated the new EnSite Precision mapping system (St. Jude Medical, Inc., St. Paul, MN, USA), which collects magnetic data for checking navigation field stability and is built on an open platform, allowing physicians to choose diagnostic and ablation catheters. We address its compatibility with RMN. To assess the clinical utility of a novel 3D mapping system (EnSite Precision mapping system) combined with RMN (Niobe ES, Stereotaxis, Inc., St. Louis, MO, USA) for atrial fibrillation (AF) ablation. In this prospective nonrandomized study, two groups of patients were treated in our center for drug refractory AF. Patients were consecutively enrolled in each group. Group A (n = 35, 14 persistent AF [PsAF]) was treated using the novel 3D mapping system combined with RMN. Group B (n = 38, 16 PsAF) was treated using Carto ® 3 (Biosense Webster, Inc., Diamond Bar, CA, USA) combined with RMN. In Group A, the left atrium (LA) was mapped with a circular magnetic catheter manually and was then replaced by a RMN ablation catheter. At the end of the procedures in Group A, the circular catheter was used for confirming field stability. In Group B, an ablation catheter was controlled by RMN to perform both LA mapping and ablation. All patients underwent pulmonary vein antrum isolation. Additional complex fractionated atrial electrograms (CFAEs) ablation was performed for PsAF. Procedural, ablation, and fluoroscopy times were recorded and complications were assessed. Electrophysiological end points were achieved in all patients. Using the novel mapping system, LA mapping time was fast (308 ± 60 seconds) with detailed anatomy points (178,831 ± 70,897) collected and magnetic points throughout LA. At the end of the procedures in Group A, the LA model was confirmed to be stable and its location was within the distance threshold (1 mm). Procedure time (117.9 ± 29.6 minutes vs. 119.2 ± 29.7 minutes, P = 0.89), fluoroscopy time (6.1 ± 2.4 minutes vs. 4.8 ± 2.2 minutes, P = 0.07), and ablation time (28.0 ± 12.9 minutes vs. 27.9 ± 15.8 minutes, P = 0.98) were similar in Group A versus Group B, respectively. No complications occurred in either group. LA mapped by the novel system is stable and reliable. Combined with RMN, it could be effectively used for AF ablation without impacting overall procedural times. © 2017 Wiley Periodicals, Inc.

  16. Patient-specific simulation in carotid artery stenting.

    PubMed

    Willaert, Willem; Aggarwal, Rajesh; Bicknell, Colin; Hamady, Mo; Darzi, Ara; Vermassen, Frank; Cheshire, Nicholas

    2010-12-01

    Patient-specific virtual reality (VR) simulation is a technologic advancement that allows planning and practice of the carotid artery stenting (CAS) procedure before it is performed on the patient. The initial findings are reported, using this novel VR technique as a tool to optimize technical and nontechnical aspects of this complex endovascular procedure. In the angiography suite, the same interventional team performed the VR rehearsal and the actual CAS on the patient. All proceedings were recorded to allow for video analysis of team, technical, and nontechnical skills. Analysis of both procedures showed identical use of endovascular tools, similar access strategy, and a high degree of similarity between the angiography images. The total procedure time (24.04 vs 60.44 minutes), fluoroscopy time (11.19 vs 21.04 minutes), and cannulation of the common carotid artery (1.35 vs 9.34) took considerably longer in reality. An extensive questionnaire revealed that all team members found that the rehearsal increased the subjective sense of teamwork (4/5), communication (4/5), and patient safety (4/5). A VR procedure rehearsal is a practical and feasible preparatory tool for CAS and shows a high correlation with the real procedure. It has the potential to enhance the technical, nontechnical, and team performance. Further research is needed to evaluate if this technology can lead to improved outcomes for patients. Copyright © 2010 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.

  17. Outcomes of transcatheter aortic valve replacement using a minimalist approach.

    PubMed

    Gurevich, Sergey; Oestreich, Brett; Kelly, Rosemary F; Mbai, Mackenzie; Bertog, Stefan; Ringsred, Karen; Lawton, Annette; Thooft, Brenda; Wagner, Julie; Garcia, Santiago

    2018-03-01

    Transcatheter Aortic Valve Replacement (TAVR) is increasingly performed using a minimalist approach under monitored anesthesia care (MAC). The safety of this approach remains controversial and adoption has been low in the US. The study cohort was comprised of 130 patients (98% male) who underwent TAVR between 4/2015 and 4/2017 at the Minneapolis VA. We compared the outcomes of 81 patients who underwent TAVR using a standard approach (standard TAVR) and 49 who underwent TAVR using a minimalist approach (minimalist TAVR). Outcome measures included device and procedural success, procedural efficiency, length of intensive care unit (ICU) and hospital stay, procedural complications and hospital readmissions. Mean age was 80 (±9) years and median (IQR) STS score was 5 (4-6). Access included transfemoral (n = 111, 85%) and alternative (n = 19, 15%). Minimalist TAVR was associated with reduced procedural time: median 101 min (IQR: 78-135) versus 127 min (IQR: 97-182, p = 0.03), fluoroscopy time: median 18 min (IQR: 12-26) versus 24 min (IQR: 16-38), p = 0.001), contrast volume 90 ml (IQR: 70-120 ml) versus 140 ml (IQR: 86-213 ml, p < 0.001). A minimalistic approach to TAVR is associated with improved procedural efficiency and reduced length of stay without compromising procedural success or clinical outcomes. Published by Elsevier Inc.

  18. Radiation Exposure and Vascular Access in Acute Coronary Syndromes: The RAD-Matrix Trial.

    PubMed

    Sciahbasi, Alessandro; Frigoli, Enrico; Sarandrea, Alessandro; Rothenbühler, Martina; Calabrò, Paolo; Lupi, Alessandro; Tomassini, Francesco; Cortese, Bernardo; Rigattieri, Stefano; Cerrato, Enrico; Zavalloni, Dennis; Zingarelli, Antonio; Calabria, Paolo; Rubartelli, Paolo; Sardella, Gennaro; Tebaldi, Matteo; Windecker, Stephan; Jüni, Peter; Heg, Dik; Valgimigli, Marco

    2017-05-23

    It remains unclear whether radial access increases the risk of operator or patient radiation exposure compared to transfemoral access when performed by expert operators. This study sought to determine whether radial access increases radiation exposure. A total of 8,404 patients, with or without ST-segment elevation acute coronary syndrome, were randomly assigned to radial or femoral access for coronary angiography and percutaneous intervention, and collected fluoroscopy time and dose-area product (DAP). RAD-MATRIX is a radiation sub-study of the MATRIX (Minimizing Adverse Haemorrhagic Events by Transradial Access Site and Systemic Implementation of AngioX) trial. We anticipated that 13 or more operators, each wearing a thorax (primary endpoint), wrist, and head (secondary endpoints) lithium fluoride thermoluminescent dosimeter, and randomizing at least 13 patients per access site, were needed to establish noninferiority of radial versus femoral access. Among 18 operators, performing 777 procedures in 767 patients, the noninferiority primary endpoint was not achieved (p value for noninferiority = 0.843). Operator equivalent dose at the thorax (77 μSv) was significantly higher with radial than femoral access (41 μSv; p = 0.02). After normalization of operator radiation dose by fluoroscopy time or DAP, the difference remained significant. Radiation dose at wrist or head did not differ between radial and femoral access. Thorax operator dose did not differ for right radial (84 μSv) compared to left radial access (52 μSv; p = 0.15). In the overall MATRIX population, fluoroscopy time and DAP were higher with radial compared to femoral access: 10 min versus 9 min (p < 0.0001) and 65 Gy·cm 2 versus 59 Gy·cm 2 (p = 0.0001), respectively. Compared to femoral access, radial access is associated with greater operator and patient radiation exposure when performed by expert operators in current practice. Radial operators and institutions should be sensitized towards radiation risks and adopt adjunctive radioprotective measures. (Minimizing Adverse Haemorrhagic Events by Transradial Access Site and Systemic Implementation of AngioX; NCT101433627). Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

  19. Professional efficiencies for diagnostic imaging services rendered by different physicians: analysis of recent medicare multiple procedure payment reduction policy.

    PubMed

    Duszak, Richard; Silva, Ezequiel; Kim, Angela J; Barr, Robert M; Donovan, William D; Kassing, Pamela; McGinty, Geraldine; Allen, Bibb

    2013-09-01

    The aim of this study was to quantify potential physician work efficiencies and appropriate multiple procedure payment reductions for different same-session diagnostic imaging studies interpreted by different physicians in the same group practice. Medicare Resource-Based Relative Value Scale data were analyzed to determine the relative contributions of various preservice, intraservice, and postservice physician diagnostic imaging work activities. An expert panel quantified potential duplications in professional work activities when separate examinations were performed during the same session by different physicians within the same group practice. Maximum potential work duplications for various imaging modalities were calculated and compared with those used as the basis of CMS payment policy. No potential intraservice work duplication was identified when different examination interpretations were rendered by different physicians in the same group practice. When multiple interpretations within the same modality were rendered by different physicians, maximum potential duplicated preservice and postservice activities ranged from 5% (radiography, fluoroscopy, and nuclear medicine) to 13.6% (CT). Maximum mean potential duplicated work relative value units ranged from 0.0049 (radiography and fluoroscopy) to 0.0413 (CT). This equates to overall potential total work reductions ranging from 1.39% (nuclear medicine) to 2.73% (CT). Across all modalities, this corresponds to maximum Medicare professional component physician fee reductions of 1.23 ± 0.38% (range, 0.95%-1.87%) for services within the same modality, much less than an order of magnitude smaller than those implemented by CMS. For services from different modalities, potential duplications were too small to quantify. Although potential efficiencies exist in physician preservice and postservice work when same-session, same-modality imaging services are rendered by different physicians in the same group practice, these are relatively minuscule and have been grossly overestimated by current CMS payment policy. Greater transparency and methodologic rigor in government payment policy development are warranted. Copyright © 2013 American College of Radiology. Published by Elsevier Inc. All rights reserved.

  20. The magnetic navigation system allows safety and high efficacy for ablation of arrhythmias

    PubMed Central

    Bauernfeind, Tamas; Akca, Ferdi; Schwagten, Bruno; de Groot, Natasja; Van Belle, Yves; Valk, Suzanne; Ujvari, Barbara; Jordaens, Luc; Szili-Torok, Tamas

    2011-01-01

    Aims We aimed to evaluate the safety and long-term efficacy of the magnetic navigation system (MNS) in a large number of patients. The MNS has the potential for improving safety and efficacy based on atraumatic catheter design and superior navigation capabilities. Methods and results In this study, 610 consecutive patients underwent ablation. Patients were divided into two age- and sex-matched groups. Ablations were performed either using MNS (group MNS, 292) or conventional manual ablation [group manual navigation (MAN), 318]. The following parameters were analysed: acute success rate, fluoroscopy time, procedure time, complications [major: pericardial tamponade, permanent atrioventricular (AV) block, major bleeding, and death; minor: minor bleeding and temporary AV block]. Recurrence rate was assessed during follow-up (15 ± 9.5 months). Subgroup analysis was performed for the following groups: atrial fibrillation, isthmus dependent and atypical atrial flutter, atrial tachycardia, AV nodal re-entrant tachycardia, circus movement tachycardia, and ventricular tachycardia (VT). Magnetic navigation system was associated with less major complications (0.34 vs. 3.2%, P = 0.01). The total numbers of complications were lower in group MNS (4.5 vs. 10%, P = 0.005). Magnetic navigation system was equally effective as MAN in acute success rate for overall groups (92 vs. 94%, P = ns). Magnetic navigation system was more successful for VTs (93 vs. 72%, P < 0.05). Less fluoroscopy was used in group MNS (30 ± 20 vs. 35 ± 25 min, P < 0.01). There were no differences in procedure times and recurrence rates for the overall groups (168 ± 67 vs. 159 ± 75 min, P = ns; 14 vs. 11%, P = ns; respectively). Conclusions Our data suggest that the use of MNS improves safety without compromising efficiency of ablations. Magnetic navigation system is more effective than manual ablation for VTs. PMID:21508006

  1. Biplane reconstruction and visualization of virtual endoscopic and fluoroscopic views for interventional device navigation

    NASA Astrophysics Data System (ADS)

    Wagner, Martin G.; Strother, Charles M.; Schafer, Sebastian; Mistretta, Charles A.

    2016-03-01

    Biplane fluoroscopic imaging is an important tool for minimally invasive procedures for the treatment of cerebrovascular diseases. However, finding a good working angle for the C-arms of the angiography system as well as navigating based on the 2D projection images can be a difficult task. The purpose of this work is to propose a novel 4D reconstruction algorithm for interventional devices from biplane fluoroscopy images and to propose new techniques for a better visualization of the results. The proposed reconstruction methods binarizes the fluoroscopic images using a dedicated noise reduction algorithm for curvilinear structures and a global thresholding approach. A topology preserving thinning algorithm is then applied and a path search algorithm minimizing the curvature of the device is used to extract the 2D device centerlines. Finally, the 3D device path is reconstructed using epipolar geometry. The point correspondences are determined by a monotonic mapping function that minimizes the reconstruction error. The three dimensional reconstruction of the device path allows the rendering of virtual fluoroscopy images from arbitrary angles as well as 3D visualizations like virtual endoscopic views or glass pipe renderings, where the vessel wall is rendered with a semi-transparent material. This work also proposes a combination of different visualization techniques in order to increase the usability and spatial orientation for the user. A combination of synchronized endoscopic and glass pipe views is proposed, where the virtual endoscopic camera position is determined based on the device tip location as well as the previous camera position using a Kalman filter in order to create a smooth path. Additionally, vessel centerlines are displayed and the path to the target is highlighted. Finally, the virtual endoscopic camera position is also visualized in the glass pipe view to further improve the spatial orientation. The proposed techniques could considerably improve the workflow of minimally invasive procedures for the treatment of cerebrovascular diseases.

  2. Prospective evaluation of the feasibility, safety, and efficacy of Cocoon Duct Occluder for transcatheter closure of large patent ductus arteriosus: A single-center study with short- and medium-term follow-up results

    PubMed Central

    Sinha, Santosh Kumar; Razi, Mahmadula; Pandey, Rama Niwas; Kumar, Prakash; Krishna, Vinay; Jha, Mukesh Jitendra; Mishra, Vikas; Asif, Mohammad; Abdali, Nasar; Tewari, Pradyot; Thakur, Ramesh; Pandey, Umeshwar; Varma, Chandra Mohan

    2017-01-01

    Objective: To evaluate the feasibility, safety, and efficacy of a novel Cocoon Duct Occluder device for the transcatheter closure (TCC) of large patent ductus arteriosus (PDA). Methods: In this prospective, non-randomized study, consecutive patients with large PDA (narrowest diameter: ≥3.5/4.0 mm in symptomatic/asymptomatic patients, respectively), who underwent TCC with Cocoon Duct Occluder at our institute between November, 2012 and June, 2016 were examined. TCC was performed using the standard technique, and devices were antegradely delivered via 6–10F delivery sheaths. Device embolization, residual shunt, hemolysis, left pulmonary artery (LPA) stenosis, procedural and fluoroscopy time, and mortality were assessed. Patients were followed-up by transthoracic echocardiography with color Doppler imaging at 24 h (D1), 1 month (D30), and 6 months (D180) after implantation. Results: A total of 57 patients (age: 11.7±2.8 years; weight: 22.3±3.5 kg) were enrolled. The mean narrowest diameter was 7.4±0.7 mm. The PDA closure was successfully performed in each patient. Fluoroscopy and procedural time was 6.7±3.2 min and 23.9±2.7 min, respectively. Postprocedural angiography revealed that 49 (85.9%) patients had immediate and complete closure, whereas 8 (14.1%) had residual shunt. Color Doppler imaging at D1 revealed complete closure in 52 (91.3%) patients. At D30, complete closure was reported in all patients and was maintained at D180. Hemolysis, embolization, obstruction of LPA or descending aorta, and death were not reported till D180. Conclusion: TCC using Cocoon Duct Occluder is feasible, safe, and effective in the management of patients with large PDA, with excellent results on short- and medium-term follow-up. PMID:29145233

  3. Computationally efficient stochastic optimization using multiple realizations

    NASA Astrophysics Data System (ADS)

    Bayer, P.; Bürger, C. M.; Finkel, M.

    2008-02-01

    The presented study is concerned with computationally efficient methods for solving stochastic optimization problems involving multiple equally probable realizations of uncertain parameters. A new and straightforward technique is introduced that is based on dynamically ordering the stack of realizations during the search procedure. The rationale is that a small number of critical realizations govern the output of a reliability-based objective function. By utilizing a problem, which is typical to designing a water supply well field, several variants of this "stack ordering" approach are tested. The results are statistically assessed, in terms of optimality and nominal reliability. This study demonstrates that the simple ordering of a given number of 500 realizations while applying an evolutionary search algorithm can save about half of the model runs without compromising the optimization procedure. More advanced variants of stack ordering can, if properly configured, save up to more than 97% of the computational effort that would be required if the entire number of realizations were considered. The findings herein are promising for similar problems of water management and reliability-based design in general, and particularly for non-convex problems that require heuristic search techniques.

  4. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ryan, E. Ronan, E-mail: ronan@ronanryan.com; Thornton, Raymond; Sofocleous, Constantinos T.

    PurposeTo quantify radiation exposure to the primary operator and staff during PET/CT-guided interventional procedures.MethodsIn this prospective study, 12 patients underwent PET/CT-guided interventions over a 6 month period. Radiation exposure was measured for the primary operator, the radiology technologist, and the nurse anesthetist by means of optically stimulated luminescence dosimeters. Radiation exposure was correlated with the procedure time and the use of in-room image guidance (CT fluoroscopy or ultrasound).ResultsThe median effective dose was 0.02 (range 0-0.13) mSv for the primary operator, 0.01 (range 0-0.05) mSv for the nurse anesthetist, and 0.02 (range 0-0.05) mSv for the radiology technologist. The median extremitymore » dose equivalent for the operator was 0.05 (range 0-0.62) mSv. Radiation exposure correlated with procedure duration and with the use of in-room image guidance. The median operator effective dose for the procedure was 0.015 mSv when conventional biopsy mode CT was used, compared to 0.06 mSv for in-room image guidance, although this did not achieve statistical significance as a result of the small sample size (p = 0.06).ConclusionThe operator dose from PET/CT-guided procedures is not significantly different than typical doses from fluoroscopically guided procedures. The major determinant of radiation exposure to the operator from PET/CT-guided interventional procedures is time spent in close proximity to the patient.« less

  5. [Surgical Correction of Scoliosis: Does Intraoperative CT Navigation Prolong Operative Time?

    PubMed

    Skála-Rosenbaum, J; Ježek, M; Džupa, V; Kadeřábek, R; Douša, P; Rusnák, R; Krbec, M

    2016-01-01

    PURPOSE OF THE STUDY The aim of the study was to compare the duration of corrective surgery for scoliosis in relation to the intra-operative use of either fluoroscopic or CT navigation. MATERIAL AND METHODS The indication for surgery was adolescent idiopathic scoliosis in younger patients and degenerative scoliosis in middleage or elderly patients. In a retrospective study, treatment outcomes in 43 consecutive patients operated on between April 2011 and April 2014 were compared. Only patients undergoing surgical correction of five or more spinal segments (fixation of six and more vertebrae) were included. RESULTS Transpedicular screw fixation of six to 13 vertebrae was performed under C-arm fluoroscopy guidance in 22 patients, and transpedicular screws were inserted in six to 14 vertebrae using the O-arm imaging system in 21 patients. A total of 246 screws were placed using the C-arm system and 340 screws were inserted using the O-arm system (p < 0.001). The procedures with use of the O-arm system were more complicated and required an average operative time longer by 48% (measured from the first skin incision to the completion of skin suture). However, the mean time needed for one screw placement (the sum of all surgical procedures with the use of a navigation technique divided by the number of screws placed using this technique) was the same in both techniques (19 min). DISCUSSION With good teamwork (surgeons, anaesthesiologists and a radiologist attending to the O-arm system), the time required to obtain one intra-operative CT scan is 3 to 5 minutes. The study showed that the mean time for placement of one screw was identical in both techniques although the average operative time was longer in surgery with O-arm navigation. The 19- minute interval was not the real placement time per screw. It was the sum of all operative times of surgical procedures (from first incision to suture completion including the whole approach within the range of planned stabilization) which used the same navigation technique divided by the number of all screws inserted during the procedures. The longer average operative time in procedures using O-arm navigation was not related to taking intra-operative O-arm scans. The authors consider surgery with an O-arm imaging system to be a safer procedure and use it currently in surgical correction of scoliosis. CONCLUSIONS The study focused on the length of surgery to correct scoliosis performed using either conventional fluoroscopy (C-arm) or intra-operative CT scanning (O-arm) showed that the mean placement time for one screw was identical in both imaging techniques when six or more vertebrae were stabilised. The use of intra-operative CT navigation did not make the surgery longer, and the higher number of inserted screws provides evidence that this technique is safer and allows us to achieve good stability of the correction procedure. Key words: virtual CT guidance, O-arm, scoliosis, transpedicular screw.

  6. The sliding rail system (monorail): description of a new technique for intravascular instrumentation and its application to coronary angioplasty.

    PubMed

    Bonzel, T; Wollschläger, H; Kasper, W; Meinertz, T; Just, H

    1987-01-01

    The sliding rail technique is a new technique for intravascular instrumentation, especially coronary stenosis dilatation. The so-called monorail balloon catheter is the first device which can be used according to this technique. The monorail catheter has a single lumen shaft and only a short central tube within the distal balloon part. With the guidewire inserted into the tube, the balloon can be advanced or retracted on the guidewire as on a sliding rail. The most relevant improvements for coronary dilatation are steerability, contrast flow and rapid and easy exchangeability of balloon catheters and other intracoronary devices. These characteristics are felt to result in a more simple and time- and fluoroscopy-saving dilatation procedure. A special transfusion catheter may also improve procedural safety. The first clinical results in 69 patients with a success rate of 96%, an emergency bypass rate of two patients (one infarction) and a stenosis improvement of 58% confirm the theoretically conceived advantages.

  7. Costing imaging procedures.

    PubMed

    Bretland, P M

    1988-01-01

    The existing National Health Service financial system makes comprehensive costing of any service very difficult. A method of costing using modern commercial methods has been devised, classifying costs into variable, semi-variable and fixed and using the principle of overhead absorption for expenditure not readily allocated to individual procedures. It proved possible to establish a cost spectrum over the financial year 1984-85. The cheapest examinations were plain radiographs outside normal working hours, followed by plain radiographs, ultrasound, special procedures, fluoroscopy, nuclear medicine, angiography and angiographic interventional procedures in normal working hours. This differs from some published figures, particularly those in the Körner report. There was some overlap between fluoroscopic interventional and the cheaper nuclear medicine procedures, and between some of the more expensive nuclear medicine procedures and the cheaper angiographic ones. Only angiographic and the few more expensive nuclear medicine procedures exceed the cost of the inpatient day. The total cost of the imaging service to the district was about 4% of total hospital expenditure. It is shown that where more procedures are undertaken, the semi-variable and fixed (including capital) elements of the cost decrease (and vice versa) so that careful study is required to assess the value of proposed economies. The method is initially time-consuming and requires a computer system with 512 Kb of memory, but once the basic costing system is established in a department, detailed financial monitoring should become practicable. The necessity for a standard comprehensive costing procedure of this nature, based on sound cost accounting principles, appears inescapable, particularly in view of its potential application to management budgeting.

  8. Cone-Beam Computed Tomography Internal Motion Tracking Should Be Used to Validate 4-Dimensional Computed Tomography for Abdominal Radiation Therapy Patients

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rankine, Leith; Wan, Hanlin; Parikh, Parag

    Purpose: To demonstrate that fiducial tracking during pretreatment Cone-Beam CT (CBCT) can accurately measure tumor motion and that this method should be used to validate 4-dimensional CT (4DCT) margins before each treatment fraction. Methods and Materials: For 31 patients with abdominal tumors and implanted fiducial markers, tumor motion was measured daily with CBCT and fluoroscopy for 202 treatment fractions. Fiducial tracking and maximum-likelihood algorithms extracted 3-dimensional fiducial trajectories from CBCT projections. The daily internal margin (IM) (ie, range of fiducial motion) was calculated for CBCT and fluoroscopy as the 5th-95th percentiles of displacement in each cardinal direction. The planning IMmore » from simulation 4DCT (IM{sub 4DCT}) was considered adequate when within ±1.2 mm (anterior–posterior, left–right) and ±3 mm (superior–inferior) of the daily measured IM. We validated CBCT fiducial tracking as an accurate predictive measure of intrafraction motion by comparing the daily measured IM{sub CBCT} with the daily IM measured by pretreatment fluoroscopy (IM{sub pre-fluoro}); these were compared with pre- and posttreatment fluoroscopy (IM{sub fluoro}) to identify those patients who could benefit from imaging during treatment. Results: Four-dimensional CT could not accurately predict intrafractional tumor motion for ≥80% of fractions in 94% (IM{sub CBCT}), 97% (IM{sub pre-fluoro}), and 100% (IM{sub fluoro}) of patients. The IM{sub CBCT} was significantly closer to IM{sub pre-fluoro} than IM{sub 4DCT} (P<.01). For patients with median treatment time t < 7.5 minutes, IM{sub CBCT} was in agreement with IM{sub fluoro} for 93% of fractions (superior–inferior), compared with 63% for the t > 7.5 minutes group, demonstrating the need for patient-specific intratreatment imaging. Conclusions: Tumor motion determined from 4DCT simulation does not accurately predict the daily motion observed on CBCT or fluoroscopy. Cone-beam CT could replace fluoroscopy for pretreatment verification of simulation IM{sub 4DCT}, reducing patient setup time and imaging dose. Patients with treatment time t > 7.5 minutes could benefit from the addition of intratreatment imaging.« less

  9. Mobile augmented reality for computer-assisted percutaneous nephrolithotomy.

    PubMed

    Müller, Michael; Rassweiler, Marie-Claire; Klein, Jan; Seitel, Alexander; Gondan, Matthias; Baumhauer, Matthias; Teber, Dogu; Rassweiler, Jens J; Meinzer, Hans-Peter; Maier-Hein, Lena

    2013-07-01

    Percutaneous nephrolithotomy (PCNL) plays an integral role in treatment of renal stones. Creating percutaneous renal access is the most important and challenging step in the procedure. To facilitate this step, we evaluated our novel mobile augmented reality (AR) system for its feasibility of use for PCNL. A tablet computer, such as an iPad[Formula: see text], is positioned above the patient with its camera pointing toward the field of intervention. The images of the tablet camera are registered with the CT image by means of fiducial markers. Structures of interest can be superimposed semi-transparently on the video images. We present a systematic evaluation by means of a phantom study. An urological trainee and two experts conducted 53 punctures on kidney phantoms. The trainee performed best with the proposed AR system in terms of puncturing time (mean: 99 s), whereas the experts performed best with fluoroscopy (mean: 59 s). iPad assistance lowered radiation exposure by a factor of 3 for the inexperienced physician and by a factor of 1.8 for the experts in comparison with fluoroscopy usage. We achieve a mean visualization accuracy of 2.5 mm. The proposed tablet computer-based AR system has proven helpful in assisting percutaneous interventions such as PCNL and shows benefits compared to other state-of-the-art assistance systems. A drawback of the system in its current state is the lack of depth information. Despite that, the simple integration into the clinical workflow highlights the potential impact of this approach to such interventions.

  10. Transcardiac endograft delivery for endovascular treatment of the ascending aorta: a feasibility study in pigs.

    PubMed

    Wipper, Sabine; Lohrenz, Christina; Ahlbrecht, Oliver; Carpenter, Sebastian W; Tsilimparis, Nikolaos; Kersten, Jan Felix; Detter, Christian; Debus, Eike S; Kölbel, Tilo

    2015-06-01

    To compare the technical feasibility and hemodynamic alterations during antegrade transcardiac access routes vs conventional transfemoral access (TFA) for endovascular treatment of the ascending aorta in a porcine model. Antegrade transseptal access (TSA), transapical access (TAA), and TFA were used for implantation of custom-made endografts into the ascending aorta under fluoroscopy (6 pigs each). Hemodynamic parameters, myocardial and cerebral blood flow, and carotid artery blood flow were evaluated during baseline (T1), sheath advancement (T2), after sheath retraction (T3), and after endograft deployment (T4). Endograft deployment was feasible in all animals; all coronary arteries remained patent. Hemodynamic parameters were comparable in all 3 study groups during all measurements. During T2, transient hemodynamic alteration occurred in all groups, with transient severe valve insufficiency in TSA and TAA reflected by the higher pulmonary to mean arterial pressure ratio (p<0.05) as compared with TFA. Values stabilized again at T3 and remained stable until T4. The innominate artery was partially occluded in 4 (TSA), 3 (TAA), and 5 (TFA) animals. There was no deterioration of myocardial or cerebral perfusion during the procedures. Endograft deployment and fluoroscopy times during TAA were shorter than in TSA and TFA. TSA, TFA, and TAA to the ascending aorta are feasible for endograft delivery to the ascending aorta in a porcine model. Transient hemodynamic instability in TSA and TAA recovered to near preoperative values. TAA appeared technically easier. © The Author(s) 2015.

  11. Percutaneous CT and Fluoroscopy-Guided Screw Fixation of Pathological Fractures in the Shoulder Girdle: Technical Report of 3 Cases

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Garnon, Julien, E-mail: juliengarnon@gmail.com; Koch, Guillaume, E-mail: Guillaume.koch@gmail.com; Ramamurthy, Nitin, E-mail: Nitin-ramamurthy@hotmail.com

    ObjectiveTo review our initial experience with percutaneous CT and fluoroscopy-guided screw fixation of pathological shoulder-girdle fractures.Materials and MethodsBetween May 2014 and June 2015, three consecutive oncologic patients (mean age 65 years; range 57–75 years) with symptomatic pathological shoulder-girdle fractures unsuitable for surgery and radiotherapy underwent percutaneous image-guided screw fixation. Fractures occurred through metastases (n = 2) or a post-ablation cavity (n = 1). Mechanical properties of osteosynthesis were adjudged superior to stand-alone cementoplasty in each case. Cannulated screws were placed under combined CT and fluoroscopic guidance with complementary radiofrequency ablation or cementoplasty to optimise local palliation and secure screw fixation, respectively, in two cases. Follow-up wasmore » undertaken every few weeks until mortality or most recent appointment.ResultsFour pathological fractures were treated in three patients (2 acromion, 1 clavicular, 1 coracoid). Mean size of associated lesion was 2.6 cm (range 1–4.5 cm). Technical success was achieved in all cases (100 %), without complications. Good palliation and restoration of mobility were observed in two cases at 2–3 months; one case could not be followed due to early post-procedural oncologic mortality.ConclusionPercutaneous image-guided shoulder-girdle osteosynthesis appears technically feasible with good short-term efficacy in this complex patient subset. Further studies are warranted to confirm these promising initial results.« less

  12. [Periprocedural and late complications after percutaneous closure of patent foramen ovale: a single centre experience].

    PubMed

    Węglarz, Przemysław; Konarska Kuszewska, Ewa; Spisak Borowska, Katarzyna; Machowski, Jerzy; Drzewiecka-Gerber, Agnieszka; Kuszewski, Piotr; Jackson, Christopher L; Opala, Grzegorz; Trusz Gluza, Maria

    2012-01-01

    Patent foramen ovale (PFO) is a potential risk factor for ischaemic stroke in young individuals. An interventional method of secondary stroke prevention in PFO patients is its percutaneous closure. To assess safety and effectiveness (i.e. lack of residual shunt) of percutaneous PFO closure in patients with history of cryptogenic cerebrovascular event. 149 patients (56 men/93 women), aged 39 ± 12 years, underwent percutaneous PFO closure. The implantation was performed under local anaesthesia, guided by trans-oesophageal echocardiography (TEE) and fluoroscopy. Follow-up trans-thoracic echocardiography (TTE) was performed at 1 month and follow-up TEE at 6-months. In cases of residual shunt, additional TEE was performed after ensuing 6 months. Effective PFO closure (no residual shunt) was achieved in 91.3% patients at 6 months and 95.3% patients at 12 months. In 2 patients transient atrial fibrillation was observed during the procedure. In 2 patients, a puncture site haematoma developed and in 1 patient superficial thrombophlebitis was noted. In 1 patient a small pericardial effusion was observed, which resolved at day 3 post-procedurally, after administration of non-steroidal anti-inflammatory drugs. Percutaneous PFO closure seems to be a safe procedure when performed in a centre with adequate expertise with regard to these procedures.

  13. Percutaneous closure of patent foramen ovale without echocardiographic guidance.

    PubMed

    Jamshidi, Peiman; Wahl, Andreas; Windecker, Stephan; Schwerzmann, Markus; Seiler, Christian; Meier, Bernhard

    2007-01-01

    A percutaneous patent foramen ovale (PFO) closure procedure includes transesophageal or intracardiac echocardiographic guidance at many centers. We investigated the feasibility and complications of the PFO closure without echocardiography. A total of 420 consecutive patients (185 women and 235 men, mean age 51 +/- 12 years) underwent percutaneous PFO closure without echocardiographic guidance using 7 different devices. Of these, 106 patients (25%) had an associated atrial septal aneurysm. The implantation was successful in 418 patients (99%). There were 12 procedural complications (3%), including embolization of the device or of parts of it with successful percutaneous removal in 5 cases, pericardial tamponade requiring pericardiocentesis in 1 patient, air embolism with transient symptoms in 3 patients, and vascular access problems in 3 patients. In none of the cases, echocardiography had to be summoned during the case or its lack was associated with acute or subsequent problems. The fluoroscopy time and procedure time were 5.4 +/- 2.7 and 25 +/- 17 minutes, respectively. Transthoracic contrast echocardiography, 24 hours after device implantation, detected a residual shunt in 19% of the patients. Percutaneous PFO closure with fluoroscopic guidance only is feasible and has low complication rates, especially with Amplatzer PFO Occluders. The added time and cost of echocardiography during the procedure is not warranted.

  14. Intraarticular Sacroiliac Joint Injection Under Computed Tomography Fluoroscopic Guidance: A Technical Note to Reduce Procedural Time and Radiation Dose

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Paik, Nam Chull, E-mail: pncspine@gmail.com

    2016-07-15

    PurposeA technique for computed tomography fluoroscopy (CTF)-guided intraarticular (IA) sacroiliac joint (SIJ) injection was devised to limit procedural time and radiation dose.MethodsOur Institutional Review Board approved this retrospective analysis and waived the requirement for informed consent. Overall, 36 consecutive diagnostic or therapeutic IA SIJ injections (unilateral, 20; bilateral, 16) performed in 34 patients (female, 18; male, 16) with a mean age of 45.5 years (range 20–76 years) under CTF guidance were analyzed, assessing technical success (i.e., IA contrast spread), procedural time, and radiation dose.ResultsAll injections were successful from a technical perspective and were free of serious complications. Respective median proceduralmore » times and effective doses of SIJ injection were as follows: unilateral, 5.28 min (range 3.58–8.00 min) and 0.11 millisievert (mSv; range 0.07–0.24 mSv); and bilateral, 6.72 min (range 4.17–21.17 min) and 0.11 mSv (range 0.09–0.51 mSv).ConclusionsGiven the high rate of technical success achieved in limited time duration and with little radiation exposure, CTF-guided IA SIJ injection is a practical and low-risk procedure.« less

  15. Antegrade Ureteral Stenting is a Good Alternative for the Retrograde Approach

    PubMed Central

    van der Meer, Rutger W.; Weltings, Saskia; van Erkel, Arian R.; Roshani, Hossain; Elzevier, Henk W.; van Dijk, Lukas C.; van Overhagen, Hans

    2017-01-01

    Background/Aims Double J (JJ) stents for treating obstructive ureteral pathology are generally inserted through a retrograde route with cystoscopic guidance. Antegrade percutaneous insertion using fluoroscopy can be performed alternatively but is less known. Indications, success rate and complications of antegrade ureteral stenting were evaluated. Methods Data of consecutive patients in which antegrade ureteral stenting was performed were retrospectively analysed using the radiology information system and patient records. Patient characteristics, details of the antegrade JJ stent insertion procedure and registered complications were collected. Furthermore, it was investigated if prior to the antegrade procedure a retrograde attempt for JJ stent insertion was performed. Results Total 130 attempts for antegrade JJ stent insertion were performed in 100 patients. A percutaneous nephrostomy catheter had already been placed in the majority of kidneys (n = 109) for initial treatment of hydronephrosis. Most prevelant indication for a JJ stent was obstructive ureteral pathology due to malignancy (n = 63). A JJ stent was successfully inserted in 125 of 130 procedures. In 21 cases, previous retrograde ureteral stenting had failed but, subsequent antegrade ureteral stenting was successful. There were 8 procedure related complications; 6 infections, 1 false tract and 1 malposition. Conclusion Antegrade percutaneous insertion of a JJ stent is a good alternative for retrograde insertion. PMID:28785193

  16. Improving Performance During Image-Guided Procedures

    PubMed Central

    Duncan, James R.; Tabriz, David

    2015-01-01

    Objective Image-guided procedures have become a mainstay of modern health care. This article reviews how human operators process imaging data and use it to plan procedures and make intraprocedural decisions. Methods A series of models from human factors research, communication theory, and organizational learning were applied to the human-machine interface that occupies the center stage during image-guided procedures. Results Together, these models suggest several opportunities for improving performance as follows: 1. Performance will depend not only on the operator’s skill but also on the knowledge embedded in the imaging technology, available tools, and existing protocols. 2. Voluntary movements consist of planning and execution phases. Performance subscores should be developed that assess quality and efficiency during each phase. For procedures involving ionizing radiation (fluoroscopy and computed tomography), radiation metrics can be used to assess performance. 3. At a basic level, these procedures consist of advancing a tool to a specific location within a patient and using the tool. Paradigms from mapping and navigation should be applied to image-guided procedures. 4. Recording the content of the imaging system allows one to reconstruct the stimulus/response cycles that occur during image-guided procedures. Conclusions When compared with traditional “open” procedures, the technology used during image-guided procedures places an imaging system and long thin tools between the operator and the patient. Taking a step back and reexamining how information flows through an imaging system and how actions are conveyed through human-machine interfaces suggest that much can be learned from studying system failures. In the same way that flight data recorders revolutionized accident investigations in aviation, much could be learned from recording video data during image-guided procedures. PMID:24921628

  17. SU-E-I-56: Diagnostic Lead Apron Radiation Exposure Comparison Between Manufacture-Stated and Measurements

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Syh, J; Patel, B; Syh, J

    2015-06-15

    Purpose: Several vendors of diagnostic lead apron used routinely in radiology/fluoroscopy claim to manufacture 0.5 mm lead equivalent shielding. The purpose of this investigation was to address the concern of the weight of lead aprons versus the radiation protection they provide. Methods: Seven diagnostic lead aprons were measured and compared for their radiation transmission and attenuation characteristics. The measurements were performed on a Philips Integris. Two settings were used, normal (76 kVp, 14.3 mA) and high (110 kVp, 12.3 mA) to represent typical patient and large patient thickness. Plastic water was placed on the table to represent patient scatter. Amore » Capintec PM-500 ion chamber was placed at approximate chest height where hospital personnel would stand. An uncovered, i.e. lead-unhindered, ion chamber reading was taken to establish the baseline reading of an unprotected personnel. The ion chamber was then wrapped with 0.5mm 99.9% pure Pb material to establish the measurement reading when a diagnostic lead apron attenuates as adequately as 0.5mm Pb. The lead aprons were measured one at a time with the ion chamber fully covered and enclosed within the aprons. Results: On Normal fluoroscopy setting, the 0.5mm pure Pb showed a transmission of 0.4%. No aprons showed a transmission value as low as 0.5mm Pb. The lowest transmission value measured from the aprons was 2.0%, having a 1.5% higher transmission than pure lead. On High fluoroscopy setting, the lowest apron transmission measurement was at 2.8%, which was comparable to the 0.5mm pure Pb which gave a transmission of 3.0%. Conclusion: At Normal fluoroscopy setting, the 0.5mm Pb provided an attenuation that could not be matched by any apron measured. At High fluoroscopy setting, only one apron exhibited comparable transmission values as 0.5mm pure Pb.« less

  18. Zero-fluoroscopy cryothermal ablation of atrioventricular nodal re-entry tachycardia guided by endovascular and endocardial catheter visualization using intracardiac echocardiography (Ice&ICE Trial).

    PubMed

    Luani, Blerim; Zrenner, Bernhard; Basho, Maksim; Genz, Conrad; Rauwolf, Thomas; Tanev, Ivan; Schmeisser, Alexander; Braun-Dullaeus, Rüdiger C

    2018-01-01

    Stochastic damage of the ionizing radiation to both patients and medical staff is a drawback of fluoroscopic guidance during catheter ablation of cardiac arrhythmias. Therefore, emerging zero-fluoroscopy catheter-guidance techniques are of great interest. We investigated, in a prospective pilot study, the feasibility and safety of the cryothermal (CA) slow-pathway ablation in patients with symptomatic atrioventricular-nodal-re-entry-tachycardia (AVNRT) using solely intracardiac echocardiography (ICE) for endovascular and endocardial catheter visualization. Twenty-five consecutive patients (mean age 55.6 ± 12.0 years, 17 female) with ECG-documentation or symptoms suggesting AVNRT underwent an electrophysiology study (EPS) in our laboratory utilizing ICE for catheter navigation. Supraventricular tachycardia was inducible in 23 (92%) patients; AVNRT was confirmed by appropriate stimulation maneuvers in 20 (80%) patients. All EPS in the AVNRT subgroup could be accomplished without need for fluoroscopy, relying solely on ICE-guidance. CA guided by anatomical location and slow-pathway potentials was successful in all patients, median cryo-mappings = 6 (IQR:3-10), median cryo-ablations = 2 (IQR:1-3). Fluoroscopy was used to facilitate the trans-septal puncture and localization of the ablation substrate in the remaining 3 patients (one focal atrial tachycardia and two atrioventricular-re-entry-tachycardias). Mean EPS duration in the AVNRT subgroup was 99.8 ± 39.6 minutes, ICE guided catheter placement 11.9 ± 5.8 minutes, time needed for diagnostic evaluation 27.1 ± 10.8 minutes, and cryo-application duration 26.3 ± 30.8 minutes. ICE-guided zero-fluoroscopy CA in AVNRT patients is feasible and safe. Real-time visualization of the true endovascular borders and cardiac structures allow for safe catheter navigation during the ICE-guided EPS and might be an alternative to visualization technologies using geometry reconstructions. © 2017 Wiley Periodicals, Inc.

  19. A comparison of CT-based navigation techniques for minimally invasive lumbar pedicle screw placement.

    PubMed

    Wood, Martin; Mannion, Richard

    2011-02-01

    A comparison of 2 surgical techniques. To determine the relative accuracy of minimally invasive lumbar pedicle screw placement using 2 different CT-based image-guided techniques. Three-dimensional intraoperative fluoroscopy systems have recently become available that provide the ability to use CT-quality images for navigation during image-guided minimally invasive spinal surgery. However, the cost of this equipment may negate any potential benefit in navigational accuracy. We therefore assess the accuracy of pedicle screw placement using an intraoperative 3-dimensional fluoroscope for guidance compared with a technique using preoperative CT images merged to intraoperative 2-dimensional fluoroscopy. Sixty-seven patients undergoing minimally invasive placement of lumbar pedicle screws (296 screws) using a navigated, image-guided technique were studied and the accuracy of pedicle screw placement assessed. Electromyography (EMG) monitoring of lumbar nerve roots was used in all. Group 1: 24 patients in whom a preoperative CT scan was merged with intraoperative 2-dimensional fluoroscopy images on the image-guidance system. Group 2: 43 patients using intraoperative 3-dimensional fluoroscopy images as the source for the image guidance system. The frequencies of pedicle breach and EMG warnings (indicating potentially unsafe screw placement) in each group were recorded. The rate of pedicle screw misplacement was 6.4% in group 1 vs 1.6% in group 2 (P=0.03). There were no cases of neurologic injury from suboptimal placement of screws. Additionally, the incidence of EMG warnings was significantly lower in group 2 (3.7% vs. 10% (P=0.03). The use of an intraoperative 3-dimensional fluoroscopy system with an image-guidance system results in greater accuracy of pedicle screw placement than the use of preoperative CT scans, although potentially dangerous placement of pedicle screws can be prevented by the use of EMG monitoring of lumbar nerve roots.

  20. Fluoroscopic evaluation of diaphragmatic motion reduction with a respiratory gated radiotherapy system.

    PubMed

    Mageras, G S; Yorke, E; Rosenzweig, K; Braban, L; Keatley, E; Ford, E; Leibel, S A; Ling, C C

    2001-01-01

    We report on initial patient studies to evaluate the performance of a commercial respiratory gating radiotherapy system. The system uses a breathing monitor, consisting of a video camera and passive infrared reflective markers placed on the patient's thorax, to synchronize radiation from a linear accelerator with the patient's breathing cycle. Six patients receiving treatment for lung cancer participated in a study of system characteristics during treatment simulation with fluoroscopy. Breathing synchronized fluoroscopy was performed initially without instruction, followed by fluoroscopy with recorded verbal instruction (i.e., when to inhale and exhale) with the tempo matched to the patient's normal breathing period. Patients tended to inhale more consistently when given instruction, as assessed by an external marker movement. This resulted in smaller variation in expiration and inspiration marker positions relative to total excursion, thereby permitting more precise gating tolerances at those parts of the breathing cycle. Breathing instruction also reduced the fraction of session times having irregular breathing as measured by the system software, thereby potentially increasing the accelerator duty factor and decreasing treatment times. Fluoroscopy studies showed external monitor movement to correlate well with that of the diaphragm in four patients, whereas time delays of up to 0.7 s in diaphragm movement were observed in two patients with impaired lung function. From fluoroscopic observations, average patient diaphragm excursion was reduced from 1.4 cm (range 0.7-2.1 cm) without gating and without breathing instruction, to 0.3 cm (range 0.2-0.5 cm) with instruction and with gating tolerances set for treatment at expiration for 25% of the breathing cycle. Patients expressed no difficulty with following instruction for the duration of a session. We conclude that the external monitor accurately predicts internal respiratory motion in most cases; however, it may be important to check with fluoroscopy for possible time delays in patients with impaired lung function. Furthermore, we observe that verbal instruction can improve breathing regularity, thus improving the performance of gated treatments with this system.

  1. Evaluation of an X-Ray Dose Profile Derived from an Optically Stimulated Luminescent Dosimeter during Computed Tomographic Fluoroscopy.

    PubMed

    Hasegawa, Hiroaki; Sato, Masanori; Tanaka, Hiroshi

    2015-01-01

    The purpose of this study was to evaluate scatter radiation dose to the subject surface during X-ray computed tomography (CT) fluoroscopy using the integrated dose ratio (IDR) of an X-ray dose profile derived from an optically stimulated luminescent (OSL) dosimeter. We aimed to obtain quantitative evidence supporting the radiation protection methods used during previous CT fluoroscopy. A multislice CT scanner was used to perform this study. OSL dosimeters were placed on the top and the lateral side of the chest phantom so that the longitudinal direction of dosimeters was parallel to the orthogonal axis-to-slice plane for measurement of dose profiles in CT fluoroscopy. Measurement of fluoroscopic conditions was performed at 120 kVp and 80 kVp. Scatter radiation dose was evaluated by calculating the integrated dose determined by OSL dosimetry. The overall percent difference of the integrated doses between OSL dosimeters and ionization chamber was 5.92%. The ratio of the integrated dose of a 100-mm length area to its tails (-50 to -6 mm, 50 to 6 mm) was the lowest on the lateral side at 80 kVp and the highest on the top at 120 kVp. The IDRs for different measurement positions were larger at 120 kVp than at 80 kVp. Similarly, the IDRs for the tube voltage between the primary X-ray beam and scatter radiation was larger on the lateral side than on the top of the phantom. IDR evaluation suggested that the scatter radiation dose has a high dependence on the position and a low dependence on tube voltage relative to the primary X-ray beam for constant dose rate fluoroscopic conditions. These results provided quantitative evidence supporting the radiation protection methods used during CT fluoroscopy in previous studies.

  2. Evaluation of an X-Ray Dose Profile Derived from an Optically Stimulated Luminescent Dosimeter during Computed Tomographic Fluoroscopy

    PubMed Central

    Hasegawa, Hiroaki; Sato, Masanori; Tanaka, Hiroshi

    2015-01-01

    The purpose of this study was to evaluate scatter radiation dose to the subject surface during X-ray computed tomography (CT) fluoroscopy using the integrated dose ratio (IDR) of an X-ray dose profile derived from an optically stimulated luminescent (OSL) dosimeter. We aimed to obtain quantitative evidence supporting the radiation protection methods used during previous CT fluoroscopy. A multislice CT scanner was used to perform this study. OSL dosimeters were placed on the top and the lateral side of the chest phantom so that the longitudinal direction of dosimeters was parallel to the orthogonal axis-to-slice plane for measurement of dose profiles in CT fluoroscopy. Measurement of fluoroscopic conditions was performed at 120 kVp and 80 kVp. Scatter radiation dose was evaluated by calculating the integrated dose determined by OSL dosimetry. The overall percent difference of the integrated doses between OSL dosimeters and ionization chamber was 5.92%. The ratio of the integrated dose of a 100-mm length area to its tails (−50 to −6 mm, 50 to 6 mm) was the lowest on the lateral side at 80 kVp and the highest on the top at 120 kVp. The IDRs for different measurement positions were larger at 120 kVp than at 80 kVp. Similarly, the IDRs for the tube voltage between the primary X-ray beam and scatter radiation was larger on the lateral side than on the top of the phantom. IDR evaluation suggested that the scatter radiation dose has a high dependence on the position and a low dependence on tube voltage relative to the primary X-ray beam for constant dose rate fluoroscopic conditions. These results provided quantitative evidence supporting the radiation protection methods used during CT fluoroscopy in previous studies. PMID:26151914

  3. The fluoroscopy time, door to balloon time, contrast volume use and prevalence of vascular access site failure with transradial versus transfemoral approach in ST segment elevation myocardial infarction: A systematic review & meta-analysis.

    PubMed

    Singh, Sukhchain; Singh, Mukesh; Grewal, Navsheen; Khosla, Sandeep

    2015-12-01

    The authors aimed to conduct first systematic review and meta-analysis in STEMI patients evaluating vascular access site failure rate, fluoroscopy time, door to balloon time and contrast volume used with transradial vs transfemoral approach (TRA vs TFA) for PCI. The PubMed, CINAHL, clinicaltrials.gov, Embase and CENTRAL databases were searched for randomized trials comparing TRA versus TFA. Random effect models were used to conduct this meta-analysis. Fourteen randomized trials comprising 3758 patients met inclusion criteria. The access site failure rate was significantly higher TRA compared to TFA (RR 3.30, CI 2.16-5.03; P=0.000). Random effect inverse variance weighted prevalence rate meta-analysis showed that access site failure rate was predicted to be 4% (95% CI 3.0-6.0%) with TRA versus 1% (95% CI 0.0-1.0 %) with TFA. Door to balloon time (Standardized mean difference [SMD] 0.30 min, 95% CI 0.23-0.37 min; P=0.000) and fluoroscopy time (Standardized mean difference 0.14 min, 95% CI 0.06-0.23 min; P=0.001) were also significantly higher in TRA. There was no difference in the amount of contrast volume used with TRA versus TFA (SMD -0.05 ml, 95% CI -0.14 to 0.04 ml; P=0.275). Statistical heterogeneity was low in cross-over rate and contrast volume use, moderate in fluoroscopy time but high in the door to balloon time comparison. Operators need to consider higher cross-over rate with TRA compared to TFA in STEMI patients while attempting PCI. Fluoroscopy and door to balloon times are negligibly higher with TRA but there is no difference in terms of contrast volume use. Copyright © 2015 Elsevier Inc. All rights reserved.

  4. Real-time x-ray fluoroscopy-based catheter detection and tracking for cardiac electrophysiology interventions

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ma Yingliang; Housden, R. James; Razavi, Reza

    2013-07-15

    Purpose: X-ray fluoroscopically guided cardiac electrophysiology (EP) procedures are commonly carried out to treat patients with arrhythmias. X-ray images have poor soft tissue contrast and, for this reason, overlay of a three-dimensional (3D) roadmap derived from preprocedural volumetric images can be used to add anatomical information. It is useful to know the position of the catheter electrodes relative to the cardiac anatomy, for example, to record ablation therapy locations during atrial fibrillation therapy. Also, the electrode positions of the coronary sinus (CS) catheter or lasso catheter can be used for road map motion correction.Methods: In this paper, the authors presentmore » a novel unified computational framework for image-based catheter detection and tracking without any user interaction. The proposed framework includes fast blob detection, shape-constrained searching and model-based detection. In addition, catheter tracking methods were designed based on the customized catheter models input from the detection method. Three real-time detection and tracking methods are derived from the computational framework to detect or track the three most common types of catheters in EP procedures: the ablation catheter, the CS catheter, and the lasso catheter. Since the proposed methods use the same blob detection method to extract key information from x-ray images, the ablation, CS, and lasso catheters can be detected and tracked simultaneously in real-time.Results: The catheter detection methods were tested on 105 different clinical fluoroscopy sequences taken from 31 clinical procedures. Two-dimensional (2D) detection errors of 0.50 {+-} 0.29, 0.92 {+-} 0.61, and 0.63 {+-} 0.45 mm as well as success rates of 99.4%, 97.2%, and 88.9% were achieved for the CS catheter, ablation catheter, and lasso catheter, respectively. With the tracking method, accuracies were increased to 0.45 {+-} 0.28, 0.64 {+-} 0.37, and 0.53 {+-} 0.38 mm and success rates increased to 100%, 99.2%, and 96.5% for the CS, ablation, and lasso catheters, respectively. Subjective clinical evaluation by three experienced electrophysiologists showed that the detection and tracking results were clinically acceptable.Conclusions: The proposed detection and tracking methods are automatic and can detect and track CS, ablation, and lasso catheters simultaneously and in real-time. The accuracy of the proposed methods is sub-mm and the methods are robust toward low-dose x-ray fluoroscopic images, which are mainly used during EP procedures to maintain low radiation dose.« less

  5. [Hybrid operating rooms: only for advanced endovascular procedures?].

    PubMed

    Verhoeven, E; Katsargyris, A; Töpel, I; Steinbauer, M

    2013-10-01

    The evolution of endovascular techniques has led to the concept of the "hybrid operating room" (hybrid OR). A hybrid OR combines the sterility of an OR in an operating theatre environment with a high-quality fixed imaging system. On the basis of these advantages it would be desirable that an angio-hybrid OR becomes a standard requirement for endovascular surgery. In Great Britain guidelines have already been published that require a hybrid OR even for normal endovascular management of the infrarenal aorta. However, in Germany there are no guidelines from professional societies or formal rules from the federal joint committee, thus in this article a classification of endovascular procedures according to their complexity and the necessary infrastructures are proposed in order to define particular procedures that should only be performed in an angio-hybrid OR. According to our experience, endovascular procedures can be classified into four categories based on their complexity and the requirements regarding fluoroscopy: level 1: standard EVAR, TEVAR, iliac and popliteal artery procedures; level 2: iliac branched (IBD) and standard (2 fenestrations for the renal arteries and a scallop for the superior mesenteric artery) fenestrated stent-grafting; level 3: more complex fenestrated procedures (three or four fenestrations); and level 4: branched stent-grafting for TAAA. At this moment it is still acceptable to perform level 1 and level 2 procedures outside of a hybrid OR. In our opinion, it is not recommended to perform level 3 and level 4 endovascular procedures without a hybrid OR. Georg Thieme Verlag KG Stuttgart · New York.

  6. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Miraglia, Roberto, E-mail: rmiraglia@ismett.edu; Maruzzelli, Luigi; Tuzzolino, Fabio

    Purpose: The aim of this study was to estimate radiation exposure in pediatric liver transplants recipients who underwent biliary interventional procedures and to compare radiation exposure levels between biliary interventional procedures performed using an image intensifier-based angiographic system (IIDS) and a flat panel detector-based interventional system (FPDS). Materials and Methods: We enrolled 34 consecutive pediatric liver transplant recipients with biliary strictures between January 2008 and March 2013 with a total of 170 image-guided procedures. The dose-area product (DAP) and fluoroscopy time was recorded for each procedure. The mean age was 61 months (range 4-192), and mean weight was 17 kgmore » (range 4-41). The procedures were classified into three categories: percutaneous transhepatic cholangiography and biliary catheter placement (n = 40); cholangiography and balloon dilatation (n = 55); and cholangiography and biliary catheter change or removal (n = 75). Ninety-two procedures were performed using an IIDS. Seventy-eight procedures performed after July 2010 were performed using an FPDS. The difference in DAP between the two angiographic systems was compared using Wilcoxon rank-sum test and a multiple linear regression model. Results: Mean DAP in the three categories was significantly greater in the group of procedures performed using the IIDS compared with those performed using the FPDS. Statistical analysis showed a p value = 0.001 for the PTBD group, p = 0.0002 for the cholangiogram and balloon dilatation group, and p = 0.00001 for the group with cholangiogram and biliary catheter change or removal. Conclusion: In our selected cohort of patients, the use of an FPDS decreases radiation exposure.« less

  7. Reference Levels for Patient Radiation Doses in Interventional Radiology: Proposed Initial Values for U.S. Practice1

    PubMed Central

    Miller, Donald L.; Kwon, Deukwoo; Bonavia, Grant H.

    2009-01-01

    Purpose: To propose initial values for patient reference levels for fluoroscopically guided procedures in the United States. Materials and Methods: This secondary analysis of data from the Radiation Doses in Interventional Radiology Procedures (RAD-IR) study was conducted under a protocol approved by the institutional review board and was HIPAA compliant. Dose distributions (percentiles) were calculated for each type of procedure in the RAD-IR study where there were data from at least 30 cases. Confidence intervals for the dose distributions were determined by using bootstrap resampling. Weight banding and size correction methods for normalizing dose to patient body habitus were tested. Results: The different methods for normalizing patient radiation dose according to patient weight gave results that were not significantly different (P > .05). The 75th percentile patient radiation doses normalized with weight banding were not significantly different from those that were uncorrected for body habitus. Proposed initial reference levels for various interventional procedures are provided for reference air kerma, kerma-area product, fluoroscopy time, and number of images. Conclusion: Sufficient data exist to permit an initial proposal of values for reference levels for interventional radiologic procedures in the United States. For ease of use, reference levels without correction for body habitus are recommended. A national registry of radiation-dose data for interventional radiologic procedures is a necessary next step to refine these reference levels. © RSNA, 2009 Supplemental material: http://radiology.rsna.org/lookup/suppl/doi:10.1148/radiol.2533090354/-/DC1 PMID:19789226

  8. [Embolization of the uterine artery in the treatment of uterine myoma].

    PubMed

    Simonetti, G; Romanini, C; Pocek, M; Piccione, E; Guazzaroni, M; Zupi, E; Gandini, R; Gabriele, A; Vaquero, E

    2001-03-01

    To propose uterine myoma embolization as an alternative to myomectomy or hysterectomy in the treatment of symptomatic myomas; to evaluate the efficacy of the procedure in terms of clinical outcome, adopting all procedural and technical precautions to ensure minimal X-ray exposure and preserve reproductive potential. Between April 1998 and February 2000, 26 patients, age range 32-54 years (mean 41 years), underwent uterine arterial embolization for menorrhagia, pelvic pain, and sensation of mass and pressure. Inclusion criteria were: single myomas, intramural localization and rich vascolarization of the lesion. Dose to patient was obtained by placing a thermoluminescent dosimeter (Harshaw, Solon, Ohio) both placed in posterior fornix of the vagina and on the skin at the beam entrance site. The procedure was performed under peridural anesthesia; polyvinil alcohol particles 355-500 mu (Contour) (Target Therapeutics, Boston Scientific Corporation, Fremont CA, USA) were employed as embolic agent. The uterine arteries were incannulated with a 5F (Glidecath, Terumo, Europe NV, Belgium) and successively 3F coaxial microcatheter (Target, Boston Scientific Corporation, Fremont CA, USA); the embolic material was injected as distally as possible. Color Power Doppler Ultrasound follow-up before and after i.v. contrast media administration (Levovist SHU 508 A, Shering, Berlin, Germany) was carried out at 15 days, at 1, at 3, at 6 months, and at 1 year from embolization. Pre-procedural evaluation and follow-up at 1 year was performed by MRI using T1 and T2 weighted images before and after Gadolinium (GdDTPA Shering, Berlin, Germany) administration. The technical success of the interventional procedure was 100% (26/26 cases). The mean fluoroscopy time was 20 minutes, and the mean number of angiographic exposures was 10. The mean estimated ovarian dose was 18.75 cGy and the mean adsorbed skin dose was 126.71 cGy. The imaging follow-up showed a 55% reduction of myoma volume at 6 months and a 75% reduction at 1 year. All patients reported a marked decrease in symptoms. No major complications were observed. The appearance of pelvic pain in the 24-48 hours after the procedure required sedation by analgesic pump; transitorial amenorrhea was observed in 3 patients. As for term complications, 2 patients have eliminated necrotic material through the vagina four weeks after procedure. The patients reported great satisfaction with the procedure. Many treatment options are currently available for symptomatic uterine myomas. One is surgical myomectomy which is associated with increased blood loss, pain and post operative morbidity and requires an additional surgical procedure for fibroma recurrence in 20-25% of patients. Another alternative treatment is hormonal therapy, which drammatically improves symptoms and reduces fibroid size although leiomyomas regrow to their original size within a few months of discontinuing treatment. Uterine embolization is a relatively new treatment for uterine fibroids that can be considered as an alternative to surgical and medical procedures. The radiation exposure adsorbed by the patient is reduced by using pulsed fluoroscopy and taking all the precautionary measures required to minimize the dose. The technical success, the patient' satisfation, the short hospitalization time and preservation of fertility confer to uterine artery embolization the role of a new alternative therapy for the treatment of symptomatic uterine myomas.

  9. A simulator for training in endovascular aneurysm repair: The use of three dimensional printers.

    PubMed

    Torres, I O; De Luccia, N

    2017-08-01

    To develop an endovascular aneurysm repair (EVAR) simulation system using three dimensional (3D) printed aneurysms, and to evaluate the impact of patient specific training prior to EVAR on the surgical performance of vascular surgery residents in a university hospital in Brazil. This was a prospective, controlled, single centre study. During 2015, the aneurysms of patients undergoing elective EVAR at São Paulo University Medical School were 3D printed and used in training sessions with vascular surgery residents. The 3D printers Stratasys-Connex 350, Formlabs-Form1+, and Makerbot were tested. Ten residents were enrolled in the control group (five residents and 30 patients in 2014) or the training group (five residents and 25 patients in 2015). The control group performed the surgery under the supervision of a senior vascular surgeon (routine procedure, without simulator training). The training group practised the surgery in a patient specific simulator prior to the routine procedure. Objective parameters were analysed, and a subjective questionnaire addressing training utility and realism was answered. Patient specific training reduced fluoroscopy time by 30% (mean 48 min, 95% confidence interval [CI] 40-58 vs. 33 min, 95% CI 26-42 [p < .01]), total procedure time by 29% (mean 292 min [95% CI 235-336] vs. 207 [95% CI 173-247]; p < .01), and volume of contrast used by 25% (mean 87 mL [95% CI 73-103] vs. 65 mL [95% CI 52-81]; p = .02). The residents considered the training useful and realistic, and reported that it increased their self confidence. The 3D printers Form1+ (using flexible resin) and Makerbot (using silicone) provided the best performance based on simulator quality and cost. An EVAR simulation system using 3D printed aneurysms was feasible. The best results were obtained with the 3D printers Form1+ (using flexible resin) and Makerbot (using silicone). Patient specific training prior to EVAR at a university hospital in Brazil improved residents' surgical performance (based on fluoroscopy time, surgery time, and volume of contrast used) and increased their self confidence. Copyright © 2017 The Authors. Published by Elsevier Ltd.. All rights reserved.

  10. Embolization of the Gastroduodenal Artery Before Selective Internal Radiotherapy: A Prospectively Randomized Trial Comparing Platinum-Fibered Microcoils with the Amplatzer Vascular Plug II

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Pech, Maciej, E-mail: maciej.pech@med.ovgu.de; Kraetsch, Annett; Wieners, Gero

    2009-05-15

    The Amplatzer Vascular Plug II (AVP II) is a novel device for transcatheter vessel occlusion, for which only limited comparative data exist. Embolotherapy of the gastroduodenal artery (GDA) is essential before internal radiotherapy (SIRT) in order to prevent radiation-induced peptic ulcerations due to migration of yttrium-90 microspheres. The purpose of this study was to compare the vascular anatomical limitations, procedure time, effectiveness, and safety of embolization of the GDA with coils versus the AVP II. Fifty patients stratified for SIRT were prospectively randomized for embolization of the GDA with either coils or the AVP II. The angle between the aortamore » and the celiac trunk, diameter of the GDA, fluoroscopy time and total time for embolization, number of embolization devices, complications, and durability of vessel occlusion at follow-up angiography for SIRT were recorded. A t-test was used for statistical analysis. Embolizations with either coils or the AVP II were technically feasible in all but two patients scheduled for embolization of the GDA with the AVP II. In both cases the plug could not be positioned due to the small celiac trunk outlet angles of 17{sup o} and 21{sup o}. The mean diameter of the GDA was 3.7 mm (range, 2.2-4.8 mm) for both groups. The procedures differed significantly in fluoroscopy time (7.8 min for coils vs. 2.6 min for the AVP II; P < 0.001) and embolization time (23.1 min for coils vs. 8.8 min for the AVP II; P < 0.001). A mean of 6.0 {+-} 3.2 coils were used for GDA embolization, while no more than one AVP II was needed for successful vessel occlusion (P < 0.001). One coil migration occurred during coil embolization, whereas no procedural complication was encountered with the use of the AVP II. Vessel reperfusion was noted in only one patient, in whom coil embolization was performed. In conclusion, embolization of the GDA with the AVP II is safe, easy, rapid, and highly effective; only an extremely sharp-angled celiac trunk outlet represented an anatomical limitation for device deployment.« less

  11. Effect of Vascular Access Site Choice on Radiation Exposure During Coronary Angiography: The REVERE Trial (Randomized Evaluation of Vascular Entry Site and Radiation Exposure).

    PubMed

    Pancholy, Samir B; Joshi, Pankaj; Shah, Sanjay; Rao, Sunil V; Bertrand, Olivier F; Patel, Tejas M

    2015-08-17

    This study sought to perform a randomized noninferiority trial of radiation exposure during cardiac catheterization comparing femoral access (FA) with left radial access (LRA) and right radial access (RRA). Increased radiation exposure with radial approach compared with femoral approach remains a controversial issue. This study randomized 1,493 patients undergoing cardiac catheterization at a tertiary care center to FA, LRA, and RRA in a 1:1:1 fashion. The primary endpoint was air kerma. The secondary endpoints included dose-area product, fluoroscopy time and operator dose per procedure, number of cineangiograms, and number of catheters. Baseline and procedural characteristics were similar among groups. No significant differences were observed in air kerma (medians: FA: 421 mGy [interquartile range (IQR): 337 to 574 mGy], LRA: 454 mGy [IQR: 331 to 643 mGy], and RRA: 483 mGy [IQR: 382 to 592 mGy], p = 0.146), dose-area product (medians: FA: 25.5 Gy cm(2) [IQR: 19.6 to 34.5 Gy cm(2)], LRA: 26.6 Gy cm(2) [IQR: 19.5 to 37.5 Gy cm(2)], and RRA: 27.7 Gy cm(2) [IQR: 21.9 to 34.4 Gy cm(2)], p = 0.40), or fluoroscopy time (medians: FA: 1.3 min [IQR: 1.0 to 1.7 min], LRA: 1.3 min [IQR: 1.0 to 1.7 min], and RRA: 1.32 min [IQR: 1.0 to 1.7 min], p = 0.19) among the 3 access sites. Median operator exposure was higher in the LRA group (3 mrem [IQR: 2 to 5 mrem], p = 0.001 vs. FA, and p = 0.0001 vs. RRA) compared with the FA (2 mrem [IQR: 2 to 4 mrem] and RRA groups (3 mrem [IQR: 2 to 5 mrem]). Radiation exposure to patients was similar during diagnostic coronary angiography with FA, RRA, and LRA. However, LRA was associated with significantly higher operator radiation exposure than were FA and RRA procedures. (Randomized Evaluation of Vascular Entry Site and Radiation Exposure [REVERE]; NCT01677481). Copyright © 2015 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

  12. New-Generation Laser-lithographed Dual-Axis Magnetically Assisted Remote-controlled Endovascular Catheter for Interventional MR Imaging: In Vitro Multiplanar Navigation at 1.5 T and 3 T versus X-ray Fluoroscopy.

    PubMed

    Moftakhar, Parham; Lillaney, Prasheel; Losey, Aaron D; Cooke, Daniel L; Martin, Alastair J; Thorne, Bradford R H; Arenson, Ronald L; Saeed, Maythem; Wilson, Mark W; Hetts, Steven W

    2015-12-01

    To assess the feasibility of multiplanar vascular navigation with a new magnetically assisted remote-controlled (MARC) catheter with real-time magnetic resonance (MR) imaging at 1.5 T and 3 T and to compare it with standard x-ray guidance in simulated endovascular catheterization procedures. A 1.6-mm-diameter custom clinical-grade microcatheter prototype with lithographed double-saddle coils at the distal tip was deflected with real-time MR imaging. Two inexperienced operators and two experienced operators catheterized anteroposterior (celiac, superior mesenteric, and inferior mesenteric arteries) and mediolateral (renal arteries) branch vessels in a cryogel abdominal aortic phantom. This was repeated with conventional x-ray fluoroscopy by using clinical catheters and guidewires. Mean procedure times and percentage success data were analyzed with linear mixed-effects regression. The MARC catheter tip was visible at 1.5 T and 3 T. Among inexperienced operators, MARC MR imaging guidance was not statistically different from x-ray guidance at 1.5 T (67% successful vessel selection turns with MR imaging vs 76% with x-ray guidance, P = .157) and at 3 T (75% successful turns with MR imaging vs 76% with x-ray guidance, P = .869). Experienced operators were more successful in catheterizing vessels with x-ray guidance (98% success within 60 seconds) than with 1.5-T (65%, P < .001) or 3-T (75%) MR imaging. Among inexperienced operators, mean procedure time was nearly equivalent by using MR imaging (31 seconds) and x-ray guidance (34 seconds, P = .436). Among experienced operators, catheterization was faster with x-ray guidance (20 seconds) compared with 1.5-T MR imaging (42 seconds, P < .001), but MARC guidance improved at 3 T (31 seconds). MARC MR imaging guidance at 3 T was not significantly different from x-ray guidance for the celiac (P = .755), superior mesenteric (P = .358), and inferior mesenteric (P = .065) arteries. Multiplanar navigation with a new MARC catheter with real-time MR imaging at 1.5 T and 3 T is feasible and comparable to x-ray guidance for anteroposterior vessels at 3 T in a vascular phantom.

  13. New-Generation Laser-lithographed Dual-Axis Magnetically Assisted Remote-controlled Endovascular Catheter for Interventional MR Imaging: In Vitro Multiplanar Navigation at 1.5 T and 3 T versus X-ray Fluoroscopy

    PubMed Central

    Moftakhar, Parham; Lillaney, Prasheel; Losey, Aaron D.; Cooke, Daniel L.; Martin, Alastair J.; Thorne, Bradford R. H.; Arenson, Ronald L.; Saeed, Maythem; Wilson, Mark W.

    2015-01-01

    Purpose To assess the feasibility of multiplanar vascular navigation with a new magnetically assisted remote-controlled (MARC) catheter with real-time magnetic resonance (MR) imaging at 1.5 T and 3 T and to compare it with standard x-ray guidance in simulated endovascular catheterization procedures. Materials and Methods A 1.6-mm–diameter custom clinical-grade microcatheter prototype with lithographed double-saddle coils at the distal tip was deflected with real-time MR imaging. Two inexperienced operators and two experienced operators catheterized anteroposterior (celiac, superior mesenteric, and inferior mesenteric arteries) and mediolateral (renal arteries) branch vessels in a cryogel abdominal aortic phantom. This was repeated with conventional x-ray fluoroscopy by using clinical catheters and guidewires. Mean procedure times and percentage success data were analyzed with linear mixed-effects regression. Results The MARC catheter tip was visible at 1.5 T and 3 T. Among inexperienced operators, MARC MR imaging guidance was not statistically different from x-ray guidance at 1.5 T (67% successful vessel selection turns with MR imaging vs 76% with x-ray guidance, P = .157) and at 3 T (75% successful turns with MR imaging vs 76% with x-ray guidance, P = .869). Experienced operators were more successful in catheterizing vessels with x-ray guidance (98% success within 60 seconds) than with 1.5-T (65%, P < .001) or 3-T (75%) MR imaging. Among inexperienced operators, mean procedure time was nearly equivalent by using MR imaging (31 seconds) and x-ray guidance (34 seconds, P = .436). Among experienced operators, catheterization was faster with x-ray guidance (20 seconds) compared with 1.5-T MR imaging (42 seconds, P < .001), but MARC guidance improved at 3 T (31 seconds). MARC MR imaging guidance at 3 T was not significantly different from x-ray guidance for the celiac (P = .755), superior mesenteric (P = .358), and inferior mesenteric (P = .065) arteries. Conclusion Multiplanar navigation with a new MARC catheter with real-time MR imaging at 1.5 T and 3 T is feasible and comparable to x-ray guidance for anteroposterior vessels at 3 T in a vascular phantom. © RSNA, 2015 Online supplemental material is available for this article. PMID:26030659

  14. Monitoring radiation use in cardiac fluoroscopy imaging procedures

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Stevens, Nathaniel T.; Steiner, Stefan H.; Smith, Ian R.

    2011-01-15

    Purpose: Timely identification of systematic changes in radiation delivery of an imaging system can lead to a reduction in risk for the patients involved. However, existing quality assurance programs involving the routine testing of equipment performance using phantoms are limited in their ability to effectively carry out this task. To address this issue, the authors propose the implementation of an ongoing monitoring process that utilizes procedural data to identify unexpected large or small radiation exposures for individual patients, as well as to detect persistent changes in the radiation output of imaging platforms. Methods: Data used in this study were obtainedmore » from records routinely collected during procedures performed in the cardiac catheterization imaging facility at St. Andrew's War Memorial Hospital, Brisbane, Australia, over the period January 2008-March 2010. A two stage monitoring process employing individual and exponentially weighted moving average (EWMA) control charts was developed and used to identify unexpectedly high or low radiation exposure levels for individual patients, as well as detect persistent changes in the radiation output delivered by the imaging systems. To increase sensitivity of the charts, we account for variation in dose area product (DAP) values due to other measured factors (patient weight, fluoroscopy time, and digital acquisition frame count) using multiple linear regression. Control charts are then constructed using the residual values from this linear regression. The proposed monitoring process was evaluated using simulation to model the performance of the process under known conditions. Results: Retrospective application of this technique to actual clinical data identified a number of cases in which the DAP result could be considered unexpected. Most of these, upon review, were attributed to data entry errors. The charts monitoring the overall system radiation output trends demonstrated changes in equipment performance associated with relocation of the equipment to a new department. When tested under simulated conditions, the EWMA chart was capable of detecting a sustained 15% increase in average radiation output within 60 cases (<1 month of operation), while a 33% increase would be signaled within 20 cases. Conclusions: This technique offers a valuable enhancement to existing quality assurance programs in radiology that rely upon the testing of equipment radiation output at discrete time frames to ensure performance security.« less

  15. The AAPM/RSNA physics tutorial for residents: digital fluoroscopy.

    PubMed

    Pooley, R A; McKinney, J M; Miller, D A

    2001-01-01

    A digital fluoroscopy system is most commonly configured as a conventional fluoroscopy system (tube, table, image intensifier, video system) in which the analog video signal is converted to and stored as digital data. Other methods of acquiring the digital data (eg, digital or charge-coupled device video and flat-panel detectors) will become more prevalent in the future. Fundamental concepts related to digital imaging in general include binary numbers, pixels, and gray levels. Digital image data allow the convenient use of several image processing techniques including last image hold, gray-scale processing, temporal frame averaging, and edge enhancement. Real-time subtraction of digital fluoroscopic images after injection of contrast material has led to widespread use of digital subtraction angiography (DSA). Additional image processing techniques used with DSA include road mapping, image fade, mask pixel shift, frame summation, and vessel size measurement. Peripheral angiography performed with an automatic moving table allows imaging of the peripheral vasculature with a single contrast material injection.

  16. Update on the recommended viewing protocol for FAXIL threshold contrast detail detectability test objects used in television fluoroscopy.

    PubMed

    Launders, J H; McArdle, S; Workman, A; Cowen, A R

    1995-01-01

    The significance of varying the viewing conditions that may affect the perceived threshold contrast of X-ray television fluoroscopy systems has been investigated. Factors investigated include the ambient room lighting and the viewing distance. The purpose of this study is to find the optimum viewing protocol with which to measure the threshold detection index. This is a particular problem when trying to compare the image quality of television fluoroscopy systems in different input field sizes. The results show that the viewing distance makes a significant difference to the perceived threshold contrast, whereas the ambient light conditions make no significant difference. Experienced observers were found to be capable of finding the optimum viewing distance for detecting details of each size, in effect using a flexible viewing distance. This allows the results from different field sizes to be normalized to account for both the magnification and the entrance air kerma rate differences, which in turn allow for a direct comparison of performance in different field sizes.

  17. Combined hysteroscopy-laparoscopy approach for excision of pelvic nitinol fragment from Essure contraceptive device: Role of intraoperative fluoroscopy for uterine conservation

    PubMed Central

    Palermo, Gianpiero D.

    2016-01-01

    We describe the successful removal of a pelvic contraceptive coil in a symptomatic 46-year-old patient who had Essure devices for four years, using a combined hysteroscopy-laparoscopy-fluoroscopy approach. Following normal hysteroscopy, at laparoscopy the right Essure implant was disrupted and its outer nitinol coil had perforated the fallopian tube. However, the inner rod (containing polyethylene terephthalate) had migrated to an extrapelvic location, near the proximal colon. In contrast, the left implant was situated within the corresponding tube. Intraoperative fluoroscopy was used to confirm complete removal of the device, which was further verified by postoperative computed tomography. The patient's condition improved after surgery and she continues to do well. This is the first report to describe this technique in managing Essure complications remote from time of insertion. Our case highlights the value and limitations of preoperative and intraoperative imaging to map Essure fragment location before surgery. PMID:27462605

  18. Real-Time 3D Fluoroscopy-Guided Large Core Needle Biopsy of Renal Masses: A Critical Early Evaluation According to the IDEAL Recommendations

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kroeze, Stephanie G. C.; Huisman, Merel; Verkooijen, Helena M.

    2012-06-15

    Introduction: Three-dimensional (3D) real-time fluoroscopy cone beam CT is a promising new technique for image-guided biopsy of solid tumors. We evaluated the technical feasibility, diagnostic accuracy, and complications of this technique for guidance of large-core needle biopsy in patients with suspicious renal masses. Methods: Thirteen patients with 13 suspicious renal masses underwent large-core needle biopsy under 3D real-time fluoroscopy cone beam CT guidance. Imaging acquisition and subsequent 3D reconstruction was done by a mobile flat-panel detector (FD) C-arm system to plan the needle path. Large-core needle biopsies were taken by the interventional radiologist. Technical success, accuracy, and safety were evaluatedmore » according to the Innovation, Development, Exploration, Assessment, Long-term study (IDEAL) recommendations. Results: Median tumor size was 2.6 (range, 1.0-14.0) cm. In ten (77%) patients, the histological diagnosis corresponded to the imaging findings: five were malignancies, five benign lesions. Technical feasibility was 77% (10/13); in three patients biopsy results were inconclusive. The lesion size of these three patients was <2.5 cm. One patient developed a minor complication. Median follow-up was 16.0 (range, 6.4-19.8) months. Conclusions: 3D real-time fluoroscopy cone beam CT-guided biopsy of renal masses is feasible and safe. However, these first results suggest that diagnostic accuracy may be limited in patients with renal masses <2.5 cm.« less

  19. Perioperative versus postoperative measurement of Taylor Spatial Frame mounting parameters.

    PubMed

    Sökücü, Sami; Demir, Bilal; Lapçin, Osman; Yavuz, Umut; Kabukçuoğlu, Yavuz S

    2014-01-01

    The aim of this study was to determine the differences, if any, between application parameters for the Taylor Spatial Frame (TSF) system obtained during surgery under fluoroscopy and after surgery from digital radiography. This retrospective study included 17 extremities of 15 patients (8 male, 7 female; mean age: 21.9 years, range: 10 to 55 years) who underwent TSF after deformity and fracture. Application parameters measured by fluoroscopy at the end of surgery after mounting the fixator were compared with parameters obtained from anteroposterior and lateral digital radiographs taken 1 day after surgery. Fixator was applied to the femur in 8 patients, tibia in 6 and radius in 3. Mean time to removal of the frame was 3.5 (range: 3 to 7) months. Mean perioperative anteroposterior, lateral and axial frame offsets of patients were 9.1 (range: 3 to 20) mm, 18.1 (range: 5 to 37) mm and 95.3 (range: 25 to 155) mm, respectively. Mean postoperative anteroposterior, lateral and axial frame offset radiographs were 11.8 (range: 2 to 30) mm, 18 (range: 6 to 47) mm and 109.5 (range: 28 to 195) mm, respectively. There was no statistically significant difference between the groups (p>0.05). While measurements taken during operation may lengthen the duration in the operation room, fluoroscopy may provide better images and is easier to perform than digital radiography. On the other hand, there is no difference between measurements taken during perioperative fluoroscopy and postoperative digital radiography.

  20. Percutaneous computer-assisted translaminar facet screw: an initial human cadaveric study.

    PubMed

    Sasso, Rick C; Best, Natalie M; Potts, Eric A

    2005-01-01

    Translaminar facet screws are a minimally invasive technique for posterior lumbar fixation with good success rates. Computer-assisted image navigation using virtual fluoroscopy allows multiple simultaneous screens in various planes to plan and drive spinal instrumentation. This study evaluates the percutaneous placement of translaminar facet screws with the use of virtual fluoroscopy as an image guidance technique. A human cadaveric study was performed with a percutaneous reference frame applied to the iliac crest. Ten translaminar facet screws were placed bilaterally at five levels. Anteroposterior and lateral images were used to navigate 4.0-mm screws through a percutaneous portal under virtual fluoroscopy. An axial computed tomographic scan through the instrumented levels was obtained after the screws were placed. Screws were graded on entry, course through the lamina, and terminus. A grading system was devised to grade the course through the lamina. All 10 screw-entry points were judged optimal at the spinous process laminar junction. There were five Grade I breeches with less than 1/2 the screw through the lamina, and five Grade 0 screw placements with the screw contained completely within the lamina. The termination point was acceptable in five screws. The screws that began on the right and terminated on the left were all found to have grade II breakouts. No screws placed the spinal canal or exiting nerve root at risk. Virtual fluoroscopy provides significant assistance in percutaneous placement of translaminar facet screws and results in safe position of entry, lamina course, and terminus.

  1. Fluoroscopic study of the normal gastrointestinal motility and measurements in the Hispaniolan Amazon parrot (Amazona ventralis).

    PubMed

    Beaufrère, Hugues; Nevarez, Javier; Taylor, W Michael; Jankowski, Gwendolyn; Rademacher, Nathalie; Gaschen, Lorrie; Pariaut, Romain; Tully, Thomas N

    2010-01-01

    Contrast fluoroscopy is a valuable tool to examine avian gastrointestinal motility. However, the lack of a standardized examination protocol and reference ranges prevents the objective interpretation of motility disorders and other gastrointestinal abnormalities. Our goals were to evaluate gastrointestinal motility in 20 Hispaniolan Amazon parrots (Amazona ventralis) by contrast fluoroscopy. Each parrot was crop-fed an equal part mixture of barium sulfate and hand-feeding formula and placed in a cardboard box for fluoroscopy. Over a 3-h period, 1.5 minute segments of lateral and ventrodorsal fluoroscopy were recorded every 30 min. The gastric cycle and patterns of intestinal motility were described. The frequency of crop contractions, esophageal boluses, and gastric cycles were determined in lateral and ventrodorsal views. A range of 3.4-6.6 gastric cycles/min was noted on the lateral view and 3.0-6.6 gastric cycles/min on the ventrodorsal view. Circular measurements of the proventriculus diameter, ventriculus width, and length were obtained using the midshaft femoral diameter as a standard reference unit. The upper limits of the reference ranges were 3.6 and 4.7 femoral units for the proventriculus diameter in the lateral and ventrodorsal view, respectively. Two consecutive measurements were obtained and the measurement technique was found to have high reproducibility. In this study, we established a standardized protocol for contrast fluoroscopic examination of the gastrointestinal tract and a reliable measurement method of the proventriculus and ventriculus using femoral units in the Hispaniolan Amazon parrot.

  2. A Modified Personalized Image-Based Drill Guide Template for Atlantoaxial Pedicle Screw Placement: A Clinical Study

    PubMed Central

    Jiang, Lianghai; Dong, Liang; Tan, Mingsheng; Qi, Yingna; Yang, Feng; Yi, Ping; Tang, Xiangsheng

    2017-01-01

    Background Atlantoaxial posterior pedicle screw fixation has been widely used for treatment of atlantoaxial instability (AAI). However, precise and safe insertion of atlantoaxial pedicle screws remains challenging. This study presents a modified drill guide template based on a previous template for atlantoaxial pedicle screw placement. Material/Methods Our study included 54 patients (34 males and 20 females) with AAI. All the patients underwent posterior atlantoaxial pedicle screw fixation: 25 patients underwent surgery with the use of a modified drill guide template (template group) and 29 patients underwent surgery via the conventional method (conventional group). In the template group, a modified drill guide template was designed for each patient. The modified drill guide template and intraoperative fluoroscopy were used for surgery in the template group, while only intraoperative fluoroscopy was used in the conventional group. Results Of the 54 patients, 52 (96.3%) completed the follow-up for more than 12 months. The template group had significantly lower intraoperative fluoroscopy frequency (p<0.001) and higher accuracy of screw insertion (p=0.045) than the conventional group. There were no significant differences in surgical duration, intraoperative blood loss, or improvement of neurological function between the 2 groups (p>0.05). Conclusions Based on the results of this study, it is feasible to use the modified drill guide template for atlantoaxial pedicle screw placement. Using the template can significantly lower the screw malposition rate and the frequency of intraoperative fluoroscopy. PMID:28301445

  3. Bimanual, intra-operative, fluoroscopy-guided removal of nasopharyngeal migratory fish bone from carotid space.

    PubMed

    Al-Abduwani, J A; Bhargava, D; Sawhney, S; Al-Abri, R

    2010-07-01

    We report a rare and unusual case of a patient with an ingested fishbone which migrated from the oropharynx to the anterior compartment of the retropharyngeal space and then to the deep neck space in the nasopharynx (i.e. the carotid space). This report aims to describe a successful, minimally invasive method of foreign body removal which avoided both major skull base surgery and any potential life-threatening complications. A secondary aim is to highlight the role of intra-operative fluoroscopy, an under-used tool. We present a 67-year-old man with a history of fish bone impaction but no fish bone visible on plain X-ray or flexible endoscopy. The diagnosis of fish bone lodged in the retropharyngeal space was confirmed by computed tomography. Surgical exploration of the anterior retropharyngeal space failed to locate the fish bone, as it had migrated to a new, unknown location. Intra-operative fluoroscopy was vital for the removal of the fish bone, as it was impossible to see with the naked eye and had migrated from its previously imaged position. The fish bone was finally retrieved bimanually using external pressure on the submandibular region, which displaced the fish bone, and fluoroscopic guidance, which assisted its removal from the nasopharyngeal lumen. To the best of our knowledge, this is the first reported case of bimanual, intra-operative, fluoroscopy-guided, intra-luminal removal of a migratory fish bone from the deep neck space in this region of the nasopharynx.

  4. 76 FR 16855 - Proposed Collection; Comment Request for Revenue Procedure 2001-21

    Federal Register 2010, 2011, 2012, 2013, 2014

    2011-03-25

    ... treat certain exchanges of debt instruments as realization events for Federal income tax purposes even... concerning Revenue Procedure 2001-21, Debt Roll-Ups. DATES: Written comments should be received on or before...: Title: Debt Roll-Ups. OMB Number: 1545-1647. Revenue Procedure Number: Revenue Procedure 2001-21...

  5. Initial clinical results with the ThermoCool® SmartTouch® Surround Flow catheter.

    PubMed

    Gonna, Hanney; Domenichini, Giulia; Zuberi, Zia; Norman, Mark; Kaba, Riyaz; Grimster, Alexander; Gallagher, Mark M

    2017-08-01

    The Biosense Webster ThermoCool® SmartTouch® Surround Flow (STSF) catheter is a recently developed ablation catheter incorporating Surround Flow (SF) technology to ensure efficient cooling and force sensing to quantify tissue contact. In our unit, it superseded the ThermoCool® SF catheter from the time of its introduction in May 2015. Procedure-related data were collected prospectively for the first 100 ablation procedures performed in our department using the STSF catheter. From a database of 654 procedures performed in our unit using the SF catheter, we selected one to match each STSF procedure, matching for procedure type, operator experience, patient age, and gender. The groups were well matched for patient age, gender, and procedure type. Procedure duration was similar in both groups (mean 225.5 vs. 221.4 min, IQR 106.5 vs. 91.5, P = 0.55), but fluoroscopy duration was shorter in the STSF group (mean 25.8 vs. 30.0, IQR 19.6 vs. 18.5, P = 0.03). No complication occurred in the STSF group. Complications occurred in two cases in the SF group (one pericardial effusion requiring drainage and one need for permanent pacing). Complete procedural success was achieved in 98 cases in the STSF group and 94 cases in the SF group (P = 0.15). The composite endpoint of procedure failure or acute complication was less common in the STSF group (2 vs. 8, P = 0.05). The STSF catheter is safe and effective in treating a range of arrhythmias. Compared with the SF catheter, it shows a trend towards improved safety-efficacy balance. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2016. For permissions please email: journals.permissions@oup.com.

  6. Radiology Physician Extenders: A Literature Review of the History and Current Roles of Physician Extenders in Medical Imaging.

    PubMed

    Sanders, Vicki L; Flanagan, Jennifer

    2015-01-01

    The purpose of the literature review was to assess the origins of radiology physician extenders and examine the current roles found in the literature of advanced practice physician extenders within medical imaging. Twenty-six articles relating to physician assistants (PAs), nurse practitioners (NPs), radiologist assistants (RAs), and nuclear medicine advanced associates (NMAAs) were reviewed to discern similarities and differences in history, scope of practice, and roles in the medical imaging field. The literature showed PAs and NPs are working mostly in interventional radiology. PAs, NPs, and RAs perform similar tasks in radiology, including history and physicals, evaluation and management, preprocedure work-up, obtaining informed consent, initial observations/reports, and post-procedure follow-up. NPs and PAs perform a variety of procedures but most commonly vascular access, paracentesis, and thoracentesis. RAs perform gastrointestinal, genitourinary, nonvascular invasive fluoroscopy procedures, and vascular access procedures. The review revealed NMAAs are working in an advanced role, but no specific performances of procedures was found in the literature, only suggested tasks and clinical competencies. PAs, NPs, and RAs are currently the three main midlevel providers used in medical imaging. These midlevel providers are being used in a variety of ways to increase the efficiency of the radiologist and provide diagnostic and therapeutic radiologic procedures to patients. NMAAs are being used in medical imaging but little literature is available on current roles in clinical practice. More research is needed to assess the exact procedures and duties being performed by these medical imaging physician extenders.

  7. RADIOPATHIES IN THE HAND OF THE SURGEON

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Marino, H.

    1962-03-01

    Several cases are described in which severe radiation injuries occurred in the hands of surgeons during operative procedures requiring fluoroscopy. The radiation lesions appeared to be aggravated by the various mechanical traumas accompanying washing-up processes preceding surgery including brushing the fingers, use of strong disinfectants and detergents, and sprinkling the hands with talcum before donning rubber gloves. These routine measures may cause acute eruptions of old radiation lesions. Various phases in the development of radiation burns on the hands, from the initial dermatitis to the late necrotic changes, are discussed. Fraquently, the early lesions are erroneously diagnosed as allergic eczemamore » so that effective therapeutic measures may not be instituted soon enough. Plastic surgical rapairs of finger lesions, in the cases described, are illustrated. (H.H.D.)« less

  8. Real-time realizations of the Bayesian Infrasonic Source Localization Method

    NASA Astrophysics Data System (ADS)

    Pinsky, V.; Arrowsmith, S.; Hofstetter, A.; Nippress, A.

    2015-12-01

    The Bayesian Infrasonic Source Localization method (BISL), introduced by Mordak et al. (2010) and upgraded by Marcillo et al. (2014) is destined for the accurate estimation of the atmospheric event origin at local, regional and global scales by the seismic and infrasonic networks and arrays. The BISL is based on probabilistic models of the source-station infrasonic signal propagation time, picking time and azimuth estimate merged with a prior knowledge about celerity distribution. It requires at each hypothetical source location, integration of the product of the corresponding source-station likelihood functions multiplied by a prior probability density function of celerity over the multivariate parameter space. The present BISL realization is generally time-consuming procedure based on numerical integration. The computational scheme proposed simplifies the target function so that integrals are taken exactly and are represented via standard functions. This makes the procedure much faster and realizable in real-time without practical loss of accuracy. The procedure executed as PYTHON-FORTRAN code demonstrates high performance on a set of the model and real data.

  9. DNA Hybridization: Nonradioactive Labeling Now Available for the Laboratory.

    ERIC Educational Resources Information Center

    Freeman, Lenore Gardner

    1984-01-01

    The advantages of DNA hybridization procedures for classroom and clinical use can now be realized by the recent development of nonradioactive DNA labeling/detection procedures. These methods (which are described) can replace the use of isotopes in standard DNA hybridization procedures. (JN)

  10. Radiation Exposure of Abdominal Cone Beam Computed Tomography

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sailer, Anna M., E-mail: anni.sailer@mumc.nl; Schurink, Geert Willem H., E-mail: gwh.schurink@mumc.nl; Wildberger, Joachim E., E-mail: j.wildberger@mumc.nl

    2015-02-15

    PurposeTo evaluate patients radiation exposure of abdominal C-arm cone beam computed tomography (CBCT).MethodsThis prospective study was approved by the institutional review board; written, informed consent was waived. Radiation exposure of abdominal CBCT was evaluated in 40 patients who underwent CBCT during endovascular interventions. Dose area product (DAP) of CBCT was documented and effective dose (ED) was estimated based on organ doses using dedicated Monte Carlo simulation software with consideration of X-ray field location and patients’ individual body weight and height. Weight-dependent ED per DAP conversion factors were calculated. CBCT radiation dose was compared to radiation dose of procedural fluoroscopy. CBCTmore » dose-related risk for cancer was assessed.ResultsMean ED of abdominal CBCT was 4.3 mSv (95 % confidence interval [CI] 3.9; 4.8 mSv, range 1.1–7.4 mSv). ED was significantly higher in the upper than in the lower abdomen (p = 0.003) and increased with patients’ weight (r = 0.55, slope = 0.045 mSv/kg, p < 0.001). Radiation exposure of CBCT corresponded to the radiation exposure of on average 7.2 fluoroscopy minutes (95 % CI 5.5; 8.8 min) in the same region of interest. Lifetime risk of exposure related cancer death was 0.033 % or less depending on age and weight.ConclusionsMean ED of abdominal CBCT was 4.3 mSv depending on X-ray field location and body weight.« less

  11. Current applications of robotics in spine surgery: a systematic review of the literature.

    PubMed

    Joseph, Jacob R; Smith, Brandon W; Liu, Xilin; Park, Paul

    2017-05-01

    OBJECTIVE Surgical robotics has demonstrated utility across the spectrum of surgery. Robotics in spine surgery, however, remains in its infancy. Here, the authors systematically review the evidence behind robotic applications in spinal instrumentation. METHODS This systematic review was conducted according to the Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA) guidelines. Relevant studies (through October 2016) that reported the use of robotics in spinal instrumentation were identified from a search of the PubMed database. Data regarding the accuracy of screw placement, surgeon learning curve, radiation exposure, and reasons for robotic failure were extracted. RESULTS Twenty-five studies describing 2 unique robots met inclusion criteria. Of these, 22 studies evaluated accuracy of spinal instrumentation. Although grading of pedicle screw accuracy was variable, the most commonly used method was the Gertzbein and Robbins system of classification. In the studies using the Gertzbein and Robbins system, accuracy (Grades A and B) ranged from 85% to 100%. Ten studies evaluated radiation exposure during the procedure. In studies that detailed fluoroscopy usage, overall fluoroscopy times ranged from 1.3 to 34 seconds per screw. Nine studies examined the learning curve for the surgeon, and 12 studies described causes of robotic failure, which included registration failure, soft-tissue hindrance, and lateral skiving of the drill guide. CONCLUSIONS Robotics in spine surgery is an emerging technology that holds promise for future applications. Surgical accuracy in instrumentation implanted using robotics appears to be high. However, the impact of robotics on radiation exposure is not clear and seems to be dependent on technique and robot type.

  12. A Comparison of Retrograde Intrarenal Surgery and Percutaneous Nephrolithotomy for Management of Renal Stones ?2 CM.

    PubMed

    Sari, Sercan; Ozok, Hakki Ugur; Cakici, Mehmet Caglar; Ozdemir, Harun; Bas, Okan; Karakoyunlu, Nihat; Sagnak, Levent; Senturk, Aykut Bugra; Ersoy, Hamit

    2017-01-18

    In this retrospective study, we aimed to compare the outcomes in patients who have been treated withpercutaneous nephrolithotomy (PNL) and retrograde intrarenal surgery (RIRS) on renal stones ≥ 2 cm size. We evaluated patients who underwent PNL or RIRS for renal stones ≥ 2 cm size betweenNovember 2011 and November 2014. Stone size, operation, fluoroscopy and hospitalization time, success rates,stone-free rates and complication rates were compared in both groups. Patients were followed for three months. 254 patients were in the PNL Group. 185 patients were in the RIRS Group. The mean age was 46.88 and48.04 years in PNL and RIRS groups, respectively.The patient and stone characteristics (age, gender, Body Mass Index, kidney anomaly, SWL history and stoneradioopacity) were similar between two groups.The mean stone size preoperatively was significantly larger in patients who were treated with PNL (26.33mm.vs24.04mm.; P = .006). In the RIRS group, the mean stone number was significantly higher than PNL group (P <.001).The mean operative, fluoroscopy and hospitalization time were significantly higher in PNL group (P < .001). Thestone-free rate was 93.3% for the PNL group and 73.5% for the RIRS group after first procedure (P < .001). Nomajor complication (Clavien III-V) occurred in the RIRS group. Although the primary treatment method for renal stones ≥ 2cm size is PNL, serious complicationscan be seen. Therefore, RIRS can be an alternative treatment option in the management of renal stones ≥2 cm size.

  13. Truly hybrid interventional MR/X-ray system: investigation of in vivo applications.

    PubMed

    Fahrig, R; Butts, K; Wen, Z; Saunders, R; Kee, S T; Sze, D Y; Daniel, B L; Laerum, F; Pelc, N J

    2001-12-01

    The purpose of this study was to provide in vivo demonstrations of the functionality of a truly hybrid interventional x-ray/magnetic resonance (MR) system. A digital flat-panel x-ray system (1,024(2) array of 200 microm pixels, 30 frames per second) was integrated into an interventional 0.5-T magnet. The hybrid system is capable of MR and x-ray imaging of the same field of view without patient movement. Two intravascular procedures were performed in a 22-kg porcine model: placement of a transjugular intrahepatic portosystemic shunt (TIPS) (x-ray-guided catheterization of the hepatic vein, MR fluoroscopy-guided portal puncture, and x-ray-guided stent placement) and mock chemoembolization (x-ray-guided subselective catheterization of a renal artery branch and MR evaluation of perfused volume). The resolution and frame rate of the x-ray fluoroscopy images were sufficient to visualize and place devices, including nitinol guidewires (0.016-0.035-inch diameter) and stents and a 2.3-F catheter. Fifth-order branches of the renal artery could be seen. The quality of both real-time (3.5 frames per second) and standard MR images was not affected by the x-ray system. During MR-guided TIPS placement, the trocar and the portal vein could be easily visualized, allowing successful puncture from hepatic to portal vein. Switching back and forth between x-ray and MR imaging modalities without requiring movement of the patient was demonstrated. The integrated nature of the system could be especially beneficial when x-ray and MR image guidance are used iteratively.

  14. Teaching basic trauma: validating FluoroSim, a digital fluoroscopic simulator for guide-wire insertion in hip surgery.

    PubMed

    Sugand, Kapil; Wescott, Robert A; Carrington, Richard; Hart, Alister; Van Duren, Bernard H

    2018-05-10

    Background and purpose - Simulation is an adjunct to surgical education. However, nothing can accurately simulate fluoroscopic procedures in orthopedic trauma. Current options for training with fluoroscopy are either intraoperative, which risks radiation, or use of expensive and unrealistic virtual reality simulators. We introduce FluoroSim, an inexpensive digital fluoroscopy simulator without the need for radiation. Patients and methods - This was a multicenter study with 26 surgeons in which everyone completed 1 attempt at inserting a guide-wire into a femoral dry bone using surgical equipment and FluoroSim. 5 objective performance metrics were recorded in real-time to assess construct validity. The surgeons were categorized based on the number of dynamic hip screws (DHS) performed: novices (< 10), intermediates (10-39) and experts (≥ 40). A 7-point Likert scale questionnaire assessed the face and content validity of FluoroSim. Results - Construct validity was present for 2 clinically validated metrics in DHS surgery. Experts and intermediates statistically significantly outperformed novices for tip-apex distance and for cut-out rate. Novices took the least number of radiographs. Face and content validity were also observed. Interpretation - FluoroSim discriminated between novice and intermediate or expert surgeons based on tip-apex distance and cut-out rate while demonstrating face and content validity. FluoroSim provides a useful adjunct to orthopedic training. Our findings concur with results from studies using other simulation modalities. FluoroSim can be implemented for education easily and cheaply away from theater in a safe and controlled environment.

  15. Radiofrequency-activated PMMA-augmentation through cannulated pedicle screws: A cadaver study to determine the biomechanical benefits in the osteoporotic spine.

    PubMed

    Karius, T; Deborre, C; Wirtz, D C; Burger, C; Prescher, A; Fölsch, A; Kabir, K; Pflugmacher, R; Goost, H

    2017-01-01

    PMMA-augmentation of pedicle screws strengthens the bone-screw-interface reducing cut-out risk. Injection of fluid cement bears a higher risk of extravasation, with difficulty of application because of inconsistent viscosity and limited injection time. To test a new method of cement augmentation of pedicle screws using radiofrequency-activated PMMA, which is suspected to be easier to apply and have less extravasations. Twenty-seven fresh-frozen human cadaver lumbar spines were divided into 18 osteoporotic (BMD ≤ 0.8 g/cm2) and 9 non-osteoporotic (BMD > 0.8 g/cm2) vertebral bodies. Bipedicular cannulated pedicle screws were implanted into the vertebral bodies; right screws were augmented with ultra-high viscosity PMMA, whereas un-cemented left pedicle screws served as negative controls. Cement distribution was controlled with fluoroscopy and CT scans. Axial pullout forces of the screws were measured with a material testing machine, and results were analyzed statistically. Fluoroscopy and CT scans showed that in all cases an adequately big cement depot with homogenous form and no signs of extravasation was injected. Pullout forces showed significant differences (p < 0.001) between the augmented and non-augmented pedicle screws for bone densities below 0.8 g/cm2 (661.9 N ± 439) and over 0.8 g/cm2 (744.9 N ± 415). Pullout-forces were significantly increased in osteoporotic as well as in non-osteoporotic vertebral bodies without a significant difference between these groups using this standardized, simple procedure with increased control and less complications like extravasation.

  16. Simulated procedure rehearsal is more effective than a preoperative generic warm-up for endovascular procedures.

    PubMed

    Willaert, Willem I M; Aggarwal, Rajesh; Daruwalla, Farhad; Van Herzeele, Isabelle; Darzi, Ara W; Vermassen, Frank E; Cheshire, Nicholas J

    2012-06-01

    Patient-specific simulated rehearsal (PsR) of a carotid artery stenting procedure (CAS) enables the interventionalist to rehearse the case before performing the procedure on the actual patient by incorporating patient-specific computed tomographic data into the simulation software. This study aimed to evaluate whether PsR of a CAS procedure can enhance the operative performance versus a virtual reality (VR) generic CAS warm-up procedure or no preparation at all. During a 10-session cognitive/technical VR course, medical residents were trained in CAS. Thereafter, in a randomized crossover study, each participant performed a patient-specific CAS case 3 times on the simulator, preceded by 3 different tasks: a PsR, a generic case, or no preparation. Technical performances were assessed using simulator-based metrics and expert-based ratings. Twenty medical residents (surgery, cardiology, radiology) were recruited. Training plateaus were observed after 10 sessions for all participants. Performances were significantly better after PsR than after a generic warm-up or no warm-up for total procedure time (16.3 ± 0.6 vs 19.7 ± 1.0 vs 20.9 ± 1.1 minutes, P = 0.001) and fluoroscopy time (9.3 ± 0.1 vs 11.2 ± 0.6 vs 11.2 ± 0.5 minutes, P = 0.022) but did not influence contrast volume or number of roadmaps used during the "real" case. PsR significantly improved the quality of performance as measured by the expert-based ratings (scores 28 vs 25 vs 25, P = 0.020). Patient-specific simulated rehearsal of a CAS procedure significantly improves operative performance, compared to a generic VR warm-up or no warm-up. This technology requires further investigation with respect to improved outcomes on patients in the clinical setting.

  17. Preoperative short hookwire placement for small pulmonary lesions: evaluation of technical success and risk factors for initial placement failure.

    PubMed

    Iguchi, Toshihiro; Hiraki, Takao; Matsui, Yusuke; Fujiwara, Hiroyasu; Masaoka, Yoshihisa; Tanaka, Takashi; Sato, Takuya; Gobara, Hideo; Toyooka, Shinichi; Kanazawa, Susumu

    2018-05-01

    To retrospectively evaluate the technical success of computed tomography fluoroscopy-guided short hookwire placement before video-assisted thoracoscopic surgery and to identify the risk factors for initial placement failure. In total, 401 short hookwire placements for 401 lesions (mean diameter 9.3 mm) were reviewed. Technical success was defined as correct positioning of the hookwire. Possible risk factors for initial placement failure (i.e., requirement for placement of an additional hookwire or to abort the attempt) were evaluated using logistic regression analysis for all procedures, and for procedures performed via the conventional route separately. Of the 401 initial placements, 383 were successful and 18 failed. Short hookwires were finally placed for 399 of 401 lesions (99.5%). Univariate logistic regression analyses revealed that in all 401 procedures only the transfissural approach was a significant independent predictor of initial placement failure (odds ratio, OR, 15.326; 95% confidence interval, CI, 5.429-43.267; p < 0.001) and for the 374 procedures performed via the conventional route only lesion size was a significant independent predictor of failure (OR 0.793, 95% CI 0.631-0.996; p = 0.046). The technical success of preoperative short hookwire placement was extremely high. The transfissural approach was a predictor initial placement failure for all procedures and small lesion size was a predictor of initial placement failure for procedures performed via the conventional route. • Technical success of preoperative short hookwire placement was extremely high. • The transfissural approach was a significant independent predictor of initial placement failure for all procedures. • Small lesion size was a significant independent predictor of initial placement failure for procedures performed via the conventional route.

  18. Gadolinium Use in Spine Pain Management Procedures for Patients with Contrast Allergies: Results in 527 Procedures

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Safriel, Yair; Ang, Roberto; Ali, Muhammed

    2008-03-15

    Introduction. To review the safety and efficacy of gadolinium in spine pain management procedures in patients at high risk for a contrast reaction and who are not suitable candidates for the use of standard non-ionic contrast. Methods. We reviewed records over a 61-month period of all image-guided spinal pain management procedures where patients had allergies making them unsuitable candidates for standard non-ionic contrast and where gadolinium was used to confirm needle tip placement prior to injection of medication. Results. Three hundred and four outpatients underwent 527 procedures. A spinal needle was used in all but 41 procedures. Gadolinium was visualizedmore » using portable C-arm fluoroscopy in vivo allowing for confirmation of needle tip location. The gadolinium dose ranged from 0.2 to 10 ml per level. The highest dose received by one patient was 15.83 ml intradiscally during a three-level discogram. Three hundred and one patients were discharged without complication or known delayed complications. One patient had documented intrathecal injection but without sequelae and 2 patients who underwent cervical procedures experienced seizures requiring admission to the intensive care unit. Both the latter patients were discharged without any further complications. Conclusion. Based on our experience we recommend using gadolinium judiciously for needle tip confirmation. We feel more confident using gadolinium in the lumbar spine and in cervical nerve blocks. Gadolinium should probably not be used as an injectate volume expander. The indications for gadolinium use in cervical needle-guided spine procedures are less clear and use of a blunt-tipped needle should be considered.« less

  19. A safety radiation marker in the cardiac catheterization lab.

    PubMed

    Kostakou, Peggy M; Damaskos, Dimitris S; Dagre, Anna G; Makavos, Georgios A; Olympios, Christophoros D

    2016-04-01

    Nowadays, in order to deal with cardiovascular disease, coronary angiography (CRA) is the best tool and gold standard for diagnosis and assessment. CRA inevitably exposes both patient and operator to radiation. The purpose of this study was to calculate the radiation exposure in association with the radiation absorbed by interventional cardiologists, in order to estimate a safety radiation marker in the catheterization laboratory. In 794 successive patients undergoing CRA and in three interventional cardiologists the following parameters were examined: radioscopy duration, radiation exposure during fluoroscopy, total radiation exposure and the number of stents per procedure. Every interventional cardiologist was exposed to 562,936 μGym2 of total radiation during CRA procedures, to 833,371 μGym2 during elective CRA + percutaneous coronary intervention (PCI) procedures and to 328,250 μGym2 during primary CRA + PCI. Hence, the total amount of radiation that every angiographer was exposed to amounted to 1,724,557.5 μGym2 (median values). During the same period, the average radiation that every angiographer absorbed was 15,253 while the average dose of radiation absorbed during one procedure was 0.06 mSv for each operator. Therefore, the ratio between radiation exposure and the radiation finally absorbed by every operator was 113:1 μGym2/mSv. The present study, indicating the ratio above, offers a safety marker in order to realistically estimate the dose absorbed by interventional cardiologists, suggesting a specified number of permitted procedures and an effective level of radiation use protection tools.

  20. Continuous Shape Estimation of Continuum Robots Using X-ray Images

    PubMed Central

    Lobaton, Edgar J.; Fu, Jinghua; Torres, Luis G.; Alterovitz, Ron

    2015-01-01

    We present a new method for continuously and accurately estimating the shape of a continuum robot during a medical procedure using a small number of X-ray projection images (e.g., radiographs or fluoroscopy images). Continuum robots have curvilinear structure, enabling them to maneuver through constrained spaces by bending around obstacles. Accurately estimating the robot’s shape continuously over time is crucial for the success of procedures that require avoidance of anatomical obstacles and sensitive tissues. Online shape estimation of a continuum robot is complicated by uncertainty in its kinematic model, movement of the robot during the procedure, noise in X-ray images, and the clinical need to minimize the number of X-ray images acquired. Our new method integrates kinematics models of the robot with data extracted from an optimally selected set of X-ray projection images. Our method represents the shape of the continuum robot over time as a deformable surface which can be described as a linear combination of time and space basis functions. We take advantage of probabilistic priors and numeric optimization to select optimal camera configurations, thus minimizing the expected shape estimation error. We evaluate our method using simulated concentric tube robot procedures and demonstrate that obtaining between 3 and 10 images from viewpoints selected by our method enables online shape estimation with errors significantly lower than using the kinematic model alone or using randomly spaced viewpoints. PMID:26279960

  1. Continuous Shape Estimation of Continuum Robots Using X-ray Images.

    PubMed

    Lobaton, Edgar J; Fu, Jinghua; Torres, Luis G; Alterovitz, Ron

    2013-05-06

    We present a new method for continuously and accurately estimating the shape of a continuum robot during a medical procedure using a small number of X-ray projection images (e.g., radiographs or fluoroscopy images). Continuum robots have curvilinear structure, enabling them to maneuver through constrained spaces by bending around obstacles. Accurately estimating the robot's shape continuously over time is crucial for the success of procedures that require avoidance of anatomical obstacles and sensitive tissues. Online shape estimation of a continuum robot is complicated by uncertainty in its kinematic model, movement of the robot during the procedure, noise in X-ray images, and the clinical need to minimize the number of X-ray images acquired. Our new method integrates kinematics models of the robot with data extracted from an optimally selected set of X-ray projection images. Our method represents the shape of the continuum robot over time as a deformable surface which can be described as a linear combination of time and space basis functions. We take advantage of probabilistic priors and numeric optimization to select optimal camera configurations, thus minimizing the expected shape estimation error. We evaluate our method using simulated concentric tube robot procedures and demonstrate that obtaining between 3 and 10 images from viewpoints selected by our method enables online shape estimation with errors significantly lower than using the kinematic model alone or using randomly spaced viewpoints.

  2. Magnetic navigation system assisted stenting of coronary bifurcation lesions.

    PubMed

    Simsek, Cihan; Magro, Michael; Patterson, Mark S; Onuma, Yoshinobu; Ciampichetti, Isabella; van Weenen, Sander; van Domburg, Ron T; Serruys, Patrick W; Boersma, Eric; van Geuns, Robert-Jan

    2011-03-01

    Magnetic guidewire assisted percutaneous coronary interventions (MPCI) could have certain advantages in coronary bifurcation lesions. We aimed to report the angiographic characteristics of the bifurcation lesions, as well as the procedural and clinical outcomes of the MPCI patients. The lesion characteristics and the treatment effect were assessed by performing diagnostic and quantitative coronary angiography with dedicated bifurcation software. A total of 76 patients (age 65 years, 78% male) were assigned to undergo MPCI, in which two-thirds of the lesions were located in LAD/D1. Fifty-seven out of 78 lesions (73%) had a diseased side branch and complex stenting techniques were used in the majority of the lesions (64%). All 59/78 (76%) fenestration attempts were successfully performed and only 13 dedicated bifurcation stents were implanted. The average acute gain in minimal luminal diameter was 1.08±0.81 mm, 0.80±0.70 mm and 0.59±0.56 mm for the proximal, distal and side branch, respectively. The procedural success was 69% with a procedure time of 107±43 minutes, fluoroscopy time of 34±24 minutes and contrast use of 338±136 ml. At a mean of 1.8-years follow-up, 15 patients (20%) had a cardiac event. MPCI is associated with good angiographic, fenestration and procedural success rates in the treatment of coronary bifurcation lesions.

  3. Toward computer-assisted image-guided congenital heart defect repair: an initial phantom analysis.

    PubMed

    Kwartowitz, David M; Mefleh, Fuad N; Baker, G Hamilton

    2017-10-01

    Radiation exposure in interventional cardiology is an important consideration, due to risk of cancer and other morbidity to the patient and clinical staff. Cardiac catheterizations rely heavily on fluoroscopic imaging exposing both patient and clinician to ionizing radiation. An image-guided surgery system capable of facilitating cardiac catheterizations was developed and tested to evaluate dose reduction. Several electromagnetically tracked tools were constructed specifically a 7-Fr catheter with five 5-degree-of-freedom magnetic seeds. Catheter guidance was accomplished using our image guidance system Kit for Navigation by Image-Focused Exploration and fluoroscopy alone. A cardiac phantom was designed and 3D printed to validate the image guidance procedure. In mock procedures, an expert clinician guided and deployed an occluder across the septal defect of the phantom heart. The image guidance method resulted in a dose of 1.26 mSv of radiation dose per procedure, while traditional guidance resulted in a dose of 3.33 mSv. Average overall dose savings for the image-guided method was nearly 2.07 mSv or 62 %. The work showed significant ([Formula: see text]) decrease in radiation dose with use of image guidance methods at the expense of a modest increase in procedure time. This study lays the groundwork for further exploration of image guidance applications in pediatric cardiology.

  4. Radiation exposure of patient and surgeon in minimally invasive kidney stone surgery.

    PubMed

    Demirci, A; Raif Karabacak, O; Yalçınkaya, F; Yiğitbaşı, O; Aktaş, C

    2016-05-01

    Percutaneous nephrolithotomy (PNL) and retrograde intrarenal surgery (RIRS) are the standard treatments used in the endoscopic treatment of kidney stones depending on the location and the size of the stone. The purpose of the study was to show the radiation exposure difference between the minimally invasive techniques by synchronously measuring the amount of radiation the patients and the surgeon received in each session, which makes our study unique. This is a prospective study which included 20 patients who underwent PNL, and 45 patients who underwent RIRS in our clinic between June 2014 and October 2014. The surgeries were assessed by dividing them into three steps: step 1: the access sheath or ureter catheter placement, step 2: lithotripsy and collection of fragments, and step 3: DJ catheter or re-entry tube insertion. For the PNL and RIRS groups, mean stone sizes were 30mm (range 16-60), and 12mm (range 7-35); mean fluoroscopy times were 337s (range 200-679), and 37s (range 7-351); and total radiation exposures were 142mBq (44.7 to 221), and 4.4mBq (0.2 to 30) respectively. Fluoroscopy times and radiation exposures at each step were found to be higher in the PNL group compared to the RIRS group. When assessed in itself, the fluoroscopy time and radiation exposure were stable in RIRS, and the radiation exposure was the highest in step 1 and the lowest in step 3 in PNL. When assessed for the 19 PNL patients and the 12 RIRS patients who had stone sizes≥2cm, the fluoroscopy time in step 1, and the radiation exposure in steps 1 and 2 were found to be higher in the PNL group than the RIRS group (P<0.001). Although there is need for more prospective randomized studies, RIRS appears to be a viable alternate for PNL because it has short fluoroscopy time and the radiation exposure is low in every step. 4. Copyright © 2016 Elsevier Masson SAS. All rights reserved.

  5. Uterine Artery Embolization for Leiomyomata: Optimization of the Radiation Dose to the Patient Using a Flat-Panel Detector Angiographic Suite

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sapoval, Marc, E-mail: marc.sapoval2@egp.aphp.fr; Pellerin, Olivier; Rehel, Jean-Luc

    The purpose of this study was to assess the ability of low-dose/low-frame fluoroscopy/angiography with a flat-panel detector angiographic suite to reduce the dose delivered to patients during uterine fibroid embolization (UFE). A two-step prospective dosimetric study was conducted, with a flat-panel detector angiography suite (Siemens Axiom Artis) integrating automatic exposure control (AEC), during 20 consecutive UFEs. Patient dosimetry was performed using calibrated thermoluminescent dosimeters placed on the lower posterior pelvis skin. The first step (10 patients; group A) consisted in UFE (bilateral embolization, calibrated microspheres) performed using the following parameters: standard fluoroscopy (15 pulses/s) and angiography (3 frames/s). The secondmore » step (next consecutive 10 patients; group B) used low-dose/low-frame fluoroscopy (7.5 pulses/s for catheterization and 3 pulses/s for embolization) and angiography (1 frame/s). We also recorded the total dose-area product (DAP) delivered to the patient and the fluoroscopy time as reported by the manufacturer's dosimetry report. The mean peak skin dose decreased from 2.4 {+-} 1.3 to 0.4 {+-} 0.3 Gy (P = 0.001) for groups A and B, respectively. The DAP values decreased from 43,113 {+-} 27,207 {mu}Gy m{sup 2} for group A to 9,515 {+-} 4,520 {mu}Gy m{sup 2} for group B (P = 0.003). The dose to ovaries and uterus decreased from 378 {+-} 238 mGy (group A) to 83 {+-} 41 mGy (group B) and from 388 {+-} 246 mGy (group A) to 85 {+-} 39 mGy (group B), respectively. Effective doses decreased from 112 {+-} 71 mSv (group A) to 24 {+-} 12 mSv (group B) (P = 0.003). In conclusion, the use of low-dose/low-frame fluoroscopy/angiography, based on a good understanding of the AEC system and also on the technique during uterine fibroid embolization, allows a significant decrease in the dose exposure to the patient.« less

  6. Reduction of Radiation Exposure Using Dynamic Trace Digital Angiography and Spot Fluoroscopy During Adrenal Venous Sampling

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Morita, Satoru, E-mail: i@imodey.com; Endo, Kenji; Suzaki, Shingo

    PurposeTo compare radiation exposure of adrenal venous sampling (AVS) using dynamic trace digital angiography (DTDA) and spot fluoroscopy with that using conventional methods.Materials and MethodsAVS was performed in 11 patients using DTDA and spot fluoroscopy (Group A) and 11 patients using conventional digital subtraction angiography (DSA) with collimation (Group B). Radiation exposure and image quality of adrenal venography using a five-point scale were compared between the groups.ResultsThe acquisition dose–area product (DAP) using DTDA and fluoro-DAP using spot fluoroscopy in Group A were lower than those using conventional DSA (5.3 ± 3.7 vs. 29.1 ± 20.1 Gy cm{sup 2}, p < 0.001) and collimation (33.3 ± 22.9 vs. 59.1 ± 35.7 Gy cm{sup 2}, p = 0.088)more » in Group B. The total DAP in Group A was significantly lower than that in Group B (38.6 ± 25.9 vs. 88.2 ± 53.6 Gy cm{sup 2}, p = 0.006). The peak skin dose for patients and operator radiation exposure in Group A were significantly lower than those in Group B (403 ± 340 vs. 771 ± 416 mGy, p = 0.030, and 17.1 ± 14.8 vs. 36.6 ± 21.7 μSv, p = 0.013). The image quality of DTDA (4.4 ± 0.6) was significantly higher than that of digital angiography (3.8 ± 0.9, p = 0.011) and equivalent to that of DSA (4.3 ± 0.8, p = 0.651).ConclusionsRadiation exposure during AVS can be reduced by approximately half for both patients and operators by using DTDA and spot fluoroscopy without sacrificing image quality.« less

  7. Fluoroscopy-free Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA) for controlling life threatening postpartum hemorrhage

    PubMed Central

    Stensaeth, Knut Haakon; Sovik, Edmund; Haig, Ingrid Natasha Ylva; Skomedal, Erna; Jorgensen, Arve

    2017-01-01

    Background Severe postpartum hemorrhage occurs in 1/1000 women giving birth. This condition is often dramatic and may be life threatening. Resuscitative Endovascular Balloon Occlusion of the Aorta (REBOA) has in recent years been introduced as a novel treatment for hemorrhagic shock. We present a series of fluoroscopy-free REBOA for controlling life threatening postpartum hemorrhage. Methods In 2008 an ‘aortic occlusion kit’ was assembled and used in three Norwegian university hospitals. The on-call interventional radiologist (IR) was to be contacted with a response time < 30 minutes in case of life threatening PPH. Demographics and characteristics were noted from the medical records. Results This retrospective study includes 36 patients treated with fluoroscopy-free REBOA for controlling severe postpartum hemorrhage in the years 2008–2015. The REBOA success rate was 100% and no patients died from REBOA related complications. Uterine artery embolization was performed in 17 (47%) patients and a hysterectomy in 16 (44%) patients. A short (11cm) introducer length was strongly associated with iliac artery thrombus formation (ρ = 0.50, P = 0.002). In addition, there was a strong negative correlation between uterine artery embolization and hysterectomy (ρ = -0.50, P = 0.002). Conclusions Our Norwegian experience indicates the clinical safety and feasibility of REBOA in life threatening PPH. Also, REBOA can be used in an emergency situation without the use of fluoroscopy with a high degree of technical success. It is important that safety implementation of REBOA is established, especially through limited aortic balloon occlusion time and a thorough balloon deflation regime. PMID:28355242

  8. Intraoperative positioning of mobile C-arms using artificial fluoroscopy

    NASA Astrophysics Data System (ADS)

    Dressel, Philipp; Wang, Lejing; Kutter, Oliver; Traub, Joerg; Heining, Sandro-Michael; Navab, Nassir

    2010-02-01

    In trauma and orthopedic surgery, imaging through X-ray fluoroscopy with C-arms is ubiquitous. This leads to an increase in ionizing radiation applied to patient and clinical staff. Placing these devices in the desired position to visualize a region of interest is a challenging task, requiring both skill of the operator and numerous X-rays for guidance. We propose an extension to C-arms for which position data is available that provides the surgeon with so called artificial fluoroscopy. This is achieved by computing digitally reconstructed radiographs (DRRs) from pre- or intraoperative CT data. The approach is based on C-arm motion estimation, for which we employ a Camera Augmented Mobile C-arm (CAMC) system, and a rigid registration of the patient to the CT data. Using this information we are able to generate DRRs and simulate fluoroscopic images. For positioning tasks, this system appears almost exactly like conventional fluoroscopy, however simulating the images from the CT data in realtime as the C-arm is moved without the application of ionizing radiation. Furthermore, preoperative planning can be done on the CT data and then visualized during positioning, e.g. defining drilling axes for pedicle approach techniques. Since our method does not require external tracking it is suitable for deployment in clinical environments and day-to-day routine. An experiment with six drillings into a lumbar spine phantom showed reproducible accuracy in positioning the C-arm, ranging from 1.1 mm to 4.1 mm deviation of marker points on the phantom compared in real and virtual images.

  9. Registration of 2D to 3D joint images using phase-based mutual information

    NASA Astrophysics Data System (ADS)

    Dalvi, Rupin; Abugharbieh, Rafeef; Pickering, Mark; Scarvell, Jennie; Smith, Paul

    2007-03-01

    Registration of two dimensional to three dimensional orthopaedic medical image data has important applications particularly in the area of image guided surgery and sports medicine. Fluoroscopy to computer tomography (CT) registration is an important case, wherein digitally reconstructed radiographs derived from the CT data are registered to the fluoroscopy data. Traditional registration metrics such as intensity-based mutual information (MI) typically work well but often suffer from gross misregistration errors when the image to be registered contains a partial view of the anatomy visible in the target image. Phase-based MI provides a robust alternative similarity measure which, in addition to possessing the general robustness and noise immunity that MI provides, also employs local phase information in the registration process which makes it less susceptible to the aforementioned errors. In this paper, we propose using the complex wavelet transform for computing image phase information and incorporating that into a phase-based MI measure for image registration. Tests on a CT volume and 6 fluoroscopy images of the knee are presented. The femur and the tibia in the CT volume were individually registered to the fluoroscopy images using intensity-based MI, gradient-based MI and phase-based MI. Errors in the coordinates of fiducials present in the bone structures were used to assess the accuracy of the different registration schemes. Quantitative results demonstrate that the performance of intensity-based MI was the worst. Gradient-based MI performed slightly better, while phase-based MI results were the best consistently producing the lowest errors.

  10. A cost effective and high fidelity fluoroscopy simulator using the Image-Guided Surgery Toolkit (IGSTK)

    NASA Astrophysics Data System (ADS)

    Gong, Ren Hui; Jenkins, Brad; Sze, Raymond W.; Yaniv, Ziv

    2014-03-01

    The skills required for obtaining informative x-ray fluoroscopy images are currently acquired while trainees provide clinical care. As a consequence, trainees and patients are exposed to higher doses of radiation. Use of simulation has the potential to reduce this radiation exposure by enabling trainees to improve their skills in a safe environment prior to treating patients. We describe a low cost, high fidelity, fluoroscopy simulation system. Our system enables operators to practice their skills using the clinical device and simulated x-rays of a virtual patient. The patient is represented using a set of temporal Computed Tomography (CT) images, corresponding to the underlying dynamic processes. Simulated x-ray images, digitally reconstructed radiographs (DRRs), are generated from the CTs using ray-casting with customizable machine specific imaging parameters. To establish the spatial relationship between the CT and the fluoroscopy device, the CT is virtually attached to a patient phantom and a web camera is used to track the phantom's pose. The camera is mounted on the fluoroscope's intensifier and the relationship between it and the x-ray source is obtained via calibration. To control image acquisition the operator moves the fluoroscope as in normal operation mode. Control of zoom, collimation and image save is done using a keypad mounted alongside the device's control panel. Implementation is based on the Image-Guided Surgery Toolkit (IGSTK), and the use of the graphics processing unit (GPU) for accelerated image generation. Our system was evaluated by 11 clinicians and was found to be sufficiently realistic for training purposes.

  11. Comparison of different approaches of estimating effective dose from reported exposure data in 3D imaging with interventional fluoroscopy systems

    NASA Astrophysics Data System (ADS)

    Svalkvist, Angelica; Hansson, Jonny; Bâth, Magnus

    2014-03-01

    Three-dimensional (3D) imaging with interventional fluoroscopy systems is today a common examination. The examination includes acquisition of two-dimensional projection images, used to reconstruct section images of the patient. The aim of the present study was to investigate the difference in resulting effective dose obtained using different levels of complexity in calculations of effective doses from these examinations. In the study the Siemens Artis Zeego interventional fluoroscopy system (Siemens Medical Solutions, Erlangen, Germany) was used. Images of anthropomorphic chest and pelvis phantoms were acquired. The exposure values obtained were used to calculate the resulting effective doses from the examinations, using the computer software PCXMC (STUK, Helsinki, Finland). The dose calculations were performed using three different methods: 1. using individual exposure values for each projection image, 2. using the mean tube voltage and the total DAP value, evenly distributed over the projection images, and 3. using the mean kV and the total DAP value, evenly distributed over smaller selection of projection images. The results revealed that the difference in resulting effective dose between the first two methods was smaller than 5%. When only a selection of projection images were used in the dose calculations the difference increased to over 10%. Given the uncertainties associated with the effective dose concept, the results indicate that dose calculations based on average exposure values distributed over a smaller selection of projection angles can provide reasonably accurate estimations of the radiation doses from 3D imaging using interventional fluoroscopy systems.

  12. Effectiveness of imaging-guided intra-articular injection: a comparison study between fluoroscopy and ultrasound.

    PubMed

    Furtado, Rita Nely Vilar; Pereira, Daniele Freitas; da Luz, Karine Rodrigues; dos Santos, Marla Francisca; Konai, Monique Sayuri; Mitraud, Sonia de Aguiar Vilela; Rosenfeld, Andre; Fernandes, Artur da Rocha Correa; Natour, Jamil

    2013-01-01

    Compare the effectiveness of ultrasound and fluoroscopy to guide intra-articular injections (IAI) in selected cases. A prospective study in our outpatient clinics at the Rheumatology Division at Universidade Federal de São Paulo (UNIFESP), Brazil, was conducted to compare the short-term (4 weeks) effectiveness of ultrasound and fluoroscopy-guided IAI in patients with rheumatic diseases. Inclusion criteria were: adults with refractory synovitis undergoing IAI with glucocorticoid. All patients had IAI performed with triamcinolone hexacetonide (20mg/ml) with varying doses according to the joint injected. A total of 71 rheumatic patients were evaluated (52 women, 44 whites). Mean age was 51.9 ± 13 years and 47 of them (66.2%) were on regular DMARD use. Analysis of the whole sample (71 patients) and hip sub-analysis (23 patients) showed that significant improvement was observed for both groups in terms of pain (P < 0.001). Global analysis also demonstrated better outcomes for patients in the FCG in terms of joint flexion (P < 0.001) and percentage change in joint flexion as compared to the USG. Likert scale score analyses demonstrated better results for the patients in the USG as compared to the FCG at the end of the study (P < 0.05). No statistically significant difference between groups was observed for any other study variable. Imaging-guided IAI improves regional pain in patients with various types of synovitis in the short term. For the vast majority of variables, no significant difference in terms of effectiveness was observed between fluoroscopy and ultrasound guided IAI.

  13. CT fluoroscopy guided transpleural cutting needle biopsy of small (≤2.5 cm) subpleural pulmonary nodules.

    PubMed

    Prosch, Helmut; Oschatz, Elisabeth; Eisenhuber, Edith; Wohlschlager, Helmut; Mostbeck, Gerhard H

    2011-01-01

    Small subpleural pulmonary lesions are difficult to biopsy. While the direct, short needle path has been reported to have a lower rate of pneumothorax, the indirect path provides a higher diagnostic yield. Therefore, we tried to optimize the needle pathway and minimize the iatrogenic pneumothorax risk by evaluating a CT fluoroscopy guided direct approach to biopsy subpleural lesions. Between 01/2005 and 01/2007, CT fluoroscopy guided core biopsies were performed in 24 patients. Using our technique, the tip of the guide needle remains outside the visceral pleura (17 G coaxial guide needle, 18 G Biopsy-gun, 15 or 22 mm needle path). The position of the lesion relative to the needle tip can be optimized using CT fluoroscopy by adjusting the breathing position of the patient. The Biopty gun is fired with the needle tip still outside the pleural space. Cytological smears are analyzed by a cytopathologist on-site, and biopsies are repeated as indicated with the coaxial needle still outside the pleura. Median nodule size was 1.6 cm (0.7-2.3 cm). A definitive diagnosis was obtained in 22 patients by histology and/or cytology. In one patient, only necrotic material could be obtained. In another patient, the intervention had to be aborted as the dyspnoic patient could not follow breathing instructions. An asymptomatic pneumothorax was present in seven patients; chest tube placement was not required. The presented biopsy approach has a high diagnostic yield and is especially advantageous for biopsies of small subpleural lesions in the lower lobes. Copyright © 2009 Elsevier Ireland Ltd. All rights reserved.

  14. Consistency of the National Realization of Dew-Point Temperature Using Standard Humidity Generators

    NASA Astrophysics Data System (ADS)

    Benyon, R.; Vicente, T.

    2012-09-01

    The comparison of two high-range standard humidity generators used by Instituto Nacional de Técnica Aeroespacial to realize dew-point temperature in the range from -10 °C to +95 °C has been performed using state-of-the art transfer standards and measurement procedures, over their overlapping range from -10 °C to +75 °C. The aim of this study is to investigate the level of agreement between the two generators, to determine any bias, and to quantify the level of consistency of the two realizations. The measurement procedures adopted to minimize the effect of the influence factors due to the transfer standards are described, and the results are discussed in the context of the declared calibration and measurement capabilities (CMCs).

  15. Efficiency of the strong satisfiability checking procedure for reactive system specifications

    NASA Astrophysics Data System (ADS)

    Shimakawa, Masaya; Hagihara, Shigeki; Yonezaki, Naoki

    2018-04-01

    Reactive systems are those that interact with their environment. To develop reactive systems without defects, it is important to describe behavior specifications in a formal language, such as linear temporal logic, and to verify the specification. Specifically, it is important to check whether specifications satisfy the property called realizability. In previous studies, we have proposed the concept of strong satisfiability as a necessary condition for realizability. Although this property of reactive system specifications is a necessary condition, many practical unrealizable specifications are also strongly unsatisfiable. Moreover, we have previously shown the theoretical complexity of the strong satisfiability problem. In the current study, we investigate the practical efficiency of the strong satisfiability checking procedure and demonstrate that strong satisfiability can be checked more efficiently than realizability.

  16. [Radiation protection in interventional cardiology].

    PubMed

    Durán, Ariel

    2015-01-01

    INTERVENTIONAL: cardiology progress makes each year a greater number of procedures and increasing complexity with a very good success rate. The problem is that this progress brings greater dose of radiation not only for the patient but to occupationally exposed workers as well. Simple methods for reducing or minimizing occupational radiation dose include: minimizing fluoroscopy time and the number of acquired images; using available patient dose reduction technologies; using good imaging-chain geometry; collimating; avoiding high-scatter areas; using protective shielding; using imaging equipment whose performance is controlled through a quality assurance programme; and wearing personal dosimeters so that you know your dose. Effective use of these methods requires both appropriate education and training in radiation protection for all interventional cardiology personnel, and the availability and use of appropriate protective tools and equipment. Regular review and investigation of personnel monitoring results, accompanied as appropriate by changes in how procedures are performed and equipment used, will ensure continual improvement in the practice of radiation protection in the interventional suite. Copyright © 2014 Instituto Nacional de Cardiología Ignacio Chávez. Published by Masson Doyma México S.A. All rights reserved.

  17. Prosthetic component segmentation with blur compensation: a fast method for 3D fluoroscopy.

    PubMed

    Tarroni, Giacomo; Tersi, Luca; Corsi, Cristiana; Stagni, Rita

    2012-06-01

    A new method for prosthetic component segmentation from fluoroscopic images is presented. The hybrid approach we propose combines diffusion filtering, region growing and level-set techniques without exploiting any a priori knowledge of the analyzed geometry. The method was evaluated on a synthetic dataset including 270 images of knee and hip prosthesis merged to real fluoroscopic data simulating different conditions of blurring and illumination gradient. The performance of the method was assessed by comparing estimated contours to references using different metrics. Results showed that the segmentation procedure is fast, accurate, independent on the operator as well as on the specific geometrical characteristics of the prosthetic component, and able to compensate for amount of blurring and illumination gradient. Importantly, the method allows a strong reduction of required user interaction time when compared to traditional segmentation techniques. Its effectiveness and robustness in different image conditions, together with simplicity and fast implementation, make this prosthetic component segmentation procedure promising and suitable for multiple clinical applications including assessment of in vivo joint kinematics in a variety of cases.

  18. The challenging image-guided abdominal mass biopsy: established and emerging techniques 'if you can see it, you can biopsy it'.

    PubMed

    Sainani, Nisha I; Arellano, Ronald S; Shyn, Paul B; Gervais, Debra A; Mueller, Peter R; Silverman, Stuart G

    2013-08-01

    Image-guided percutaneous biopsy of abdominal masses is among the most commonly performed procedures in interventional radiology. While most abdominal masses are readily amenable to percutaneous biopsy, some may be technically challenging for a number of reasons. Low lesion conspicuity, small size, overlying or intervening structures, motion, such as that due to respiration, are some of the factors that can influence the ability and ultimately the success of an abdominal biopsy. Various techniques or technologies, such as choice of imaging modality, use of intravenous contrast and anatomic landmarks, patient positioning, organ displacement or trans-organ approach, angling CT gantry, triangulation method, real-time guidance with CT fluoroscopy or ultrasound, sedation or breath-hold, pre-procedural image fusion, electromagnetic tracking, and others, when used singularly or in combination, can overcome these challenges to facilitate needle placement in abdominal masses that otherwise would be considered not amenable to percutaneous biopsy. Familiarity and awareness of these techniques allows the interventional radiologist to expand the use of percutaneous biopsy in clinical practice, and help choose the most appropriate technique for a particular patient.

  19. Medical health physics: a review.

    PubMed

    Vetter, Richard J

    2004-05-01

    Medical health physics is the profession dedicated to the protection of healthcare providers, members of the public, and patients from unwarranted radiation exposure. Medical health physicists must be knowledgeable in the principles of health physics and in the applications of radiation in medicine. Advances in medical health physics require the definition of problems, testing of hypotheses, and gathering of evidence to defend changes in health physics practice and to assist medical practitioners in making changes in their practices as appropriate. Advances in radiation medicine have resulted in new modalities and procedures, some of which have significant potential to cause serious harm. Examples included in this review include radiologic procedures that require very long fluoroscopy times, radiolabeled monoclonal antibodies, and intravascular brachytherapy. This review summarizes evidence that supports changes in consensus recommendations, regulations, and health physics practices associated with recent advances in radiology, nuclear medicine, and radiation oncology. Medical health physicists must continue to gather evidence to support intelligent but practical methods for protection of personnel, the public, and patients as modalities and applications evolve in the practice of medicine.

  20. Medical health physics: a review.

    PubMed

    Vetter, Richard J

    2005-06-01

    Medical health physics is the profession dedicated to the protection of healthcare providers, members of the public, and patients from unwarranted radiation exposure. Medical health physicists must be knowledgeable in the principles of health physics and in the applications of radiation in medicine. Advances in medical health physics require the definition of problems, testing of hypotheses, and gathering of evidence to defend changes in health physics practice and to assist medical practitioners in making changes in their practices as appropriate. Advances in radiation medicine have resulted in new modalities and procedures, some of which have significant potential to cause serious harm. Examples included in this review include radiologic procedures that require very long fluoroscopy times, radiolabeled monoclonal antibodies, and intravascular brachytherapy. This review summarizes evidence that supports changes in consensus recommendations, regulations, and health physics practices associated with recent advances in radiology, nuclear medicine, and radiation oncology. Medical health physicists must continue to gather evidence to support intelligent but practical methods for protection of personnel, the public, and patients as modalities and applications evolve in the practice of medicine.

  1. Fluoroless catheter ablation of various right and left sided supra-ventricular tachycardias in children and adolescents.

    PubMed

    Jan, Matevž; Žižek, David; Rupar, Katja; Mazić, Uroš; Kuhelj, Dimitrij; Lakič, Nikola; Geršak, Borut

    2016-11-01

    Electrophysiology study (EPS) and catheter ablation (CA) in children and adolescents carries a potentially harmful effect of radiation exposure when performed with the use of fluoroscopy. Our aim was to evaluate the feasibility, safety and effectiveness of fluoroless EPS and CA of various supra-ventricular tachycardias (SVTs) with the use of the 3D mapping system and intracardiac echocardiography (ICE). Forty-three consecutive children and adolescents (age 13 ± 3 years) underwent fluoroless EPS and CA for various supra-ventricular tachycardias. A three-dimensional (3D) mapping system NavX™ was used for guidance of diagnostic and ablation catheters in the heart. ICE was used as a fundamental imaging tool for transseptal punctures. Acute procedural success rate was 100 %. There were no procedure related complications and short-term follow up (10 ± 3 months) revealed 93 % arrhythmia free survival rate. Fluoroless CA of various SVTs in the paediatric population is feasible, safe and can be performed successfully with 3D mapping system and ICE.

  2. AN ANALYSIS OF OPERATING PHYSICIAN AND PATIENT RADIATION EXPOSURE DURING RADIAL CORONARY ANGIOPLASTIES.

    PubMed

    Tarighatnia, Ali; Mesbahi, Asghar; Alian, Amir Hossein Mohammad; Koleini, Evin; Nader, Nader

    2018-03-23

    The objective of this study was to evaluate radiation exposure levels in conjunction with operator dose implemented, patient vascular characteristics, and other technical angiographic parameters. In total, 756 radial coronary angioplasties were evaluated in a major metropolitan general hospital in Tabriz, Iran. The classification of coronary lesions was based on the ACC/AHA system. One interventional cardiologist performed all of the procedures using a single angiography unit. The mean kerma-area product and mean cumulative dose for all cases was 5081 μGy m2 and 814.44 mGy, respectively. Average times of 26.16 and 9.1 min were recorded for the overall procedure and fluoroscopy, respectively. A strong correlation was demonstrated between types of lesions, number of stents and vessels treated in relation to physician radiation exposure. It was determined that operator radiation exposure levels for percutaneous coronary interventions lesions (complex) were higher than that of simple and moderate lesions. In addition, operator radiation exposure levels increased with the treatment of more coronary vessels and implementation of additional stents.

  3. Retroperitoneal hemorrhage from an unrecognized puncture of the lumbar right segmental artery during lumbar chemical sympathectomy: diagnosis and management.

    PubMed

    Shin, Ho-Jin; Choi, Yun-Mi; Kim, Hye-Jin; Lee, Sun-Jae; Yoon, Seok-Hyun; Kim, Kyung-Hoon

    2014-12-01

    Lumbar chemical sympathectomy has been performed using fluoroscopic guidance for needle positioning. An 84 year old woman with atherosclerosis obliterans was referred to the pain clinic for intractable cold allodynia of her right foot. A thermogram showed decreased temperature of both feet compared with temperatures above both ankles. The patient agreed to undergo lumbar chemical sympathectomy using fluoroscopy after being informed of the associated risks of nerve injury, hemorrhage, infection, transient back pain, and transient hypotension. During the procedure and three hours afterward, no abnormal signs or symptoms were found except an increase in right leg temperature. The patient was ambulatory after the procedure. However, one day after undergoing lumbar chemical sympathectomy, she visited our emergency department for abdominal discomfort and postural dizziness. Her blood pressure was 80/50 mmHg, and flank tenderness was noted. Retroperitoneal hemorrhage from the second right lumbar segmental artery was shown on computed tomography and angiography. Vital signs were stabilized immediately after embolization into the right lumbar segmental artery. Copyright © 2014 Elsevier Inc. All rights reserved.

  4. Robust model-based 3d/3D fusion using sparse matching for minimally invasive surgery.

    PubMed

    Neumann, Dominik; Grbic, Sasa; John, Matthias; Navab, Nassir; Hornegger, Joachim; Ionasec, Razvan

    2013-01-01

    Classical surgery is being disrupted by minimally invasive and transcatheter procedures. As there is no direct view or access to the affected anatomy, advanced imaging techniques such as 3D C-arm CT and C-arm fluoroscopy are routinely used for intra-operative guidance. However, intra-operative modalities have limited image quality of the soft tissue and a reliable assessment of the cardiac anatomy can only be made by injecting contrast agent, which is harmful to the patient and requires complex acquisition protocols. We propose a novel sparse matching approach for fusing high quality pre-operative CT and non-contrasted, non-gated intra-operative C-arm CT by utilizing robust machine learning and numerical optimization techniques. Thus, high-quality patient-specific models can be extracted from the pre-operative CT and mapped to the intra-operative imaging environment to guide minimally invasive procedures. Extensive quantitative experiments demonstrate that our model-based fusion approach has an average execution time of 2.9 s, while the accuracy lies within expert user confidence intervals.

  5. Augmented reality fluoroscopy simulation of the guide-wire insertion in DHS surgery: A proof of concept study.

    PubMed

    van Duren, B H; Sugand, K; Wescott, R; Carrington, R; Hart, A

    2018-05-01

    Hip fractures contribute to a significant clinical burden globally with over 1.6 million cases per annum and up to 30% mortality rate within the first year. Insertion of a dynamic hip screw (DHS) is a frequently performed procedure to treat extracapsular neck of femur fractures. Poorly performed DHS fixation of extracapsular neck of femur fractures can result in poor mobilisation, chronic pain, and increased cut-out rate requiring revision surgery. A realistic, affordable, and portable fluoroscopic simulation system can improve performance metrics in trainees, including the tip-apex distance (the only clinically validated outcome), and improve outcomes. We developed a digital fluoroscopic imaging simulator using orthogonal cameras to track coloured markers attached to the guide-wire which created a virtual overlay on fluoroscopic images of the hip. To test the accuracy with which the augmented reality system could track a guide-wire, a standard workshop femur was used to calibrate the system with a positional marker fixed to indicate the apex; this allowed for comparison between guide-wire tip-apex distance (TAD) calculated by the system to be compared to that physically measured. Tests were undertaken to determine: (1) how well the apex could be targeted; (2) the accuracy of the calculated TAD. (3) The number of iterations through the algorithm giving the optimal accuracy-time relationship. The calculated TAD was found to have an average root mean square error of 4.2 mm. The accuracy of the algorithm was shown to increase with the number of iterations up to 20 beyond which the error asymptotically converged to an error of 2 mm. This work demonstrates a novel augmented reality simulation of guide-wire insertion in DHS surgery. To our knowledge this has not been previously achieved. In contrast to virtual reality, augmented reality is able to simulate fluoroscopy while allowing the trainee to interact with real instrumentation and performing the procedure on workshop bone models. Copyright © 2018 IPEM. Published by Elsevier Ltd. All rights reserved.

  6. Single freeze per vein strategy with the second-generation cryoballoon for atrial fibrillation: a propensity score-matched study between 180- and 240-s application time in a large cohort of patients.

    PubMed

    De Regibus, Valentina; Abugattas, Juan-Pablo; Iacopino, Saverio; Mugnai, Giacomo; Storti, Cesare; Conte, Giulio; Auricchio, Angelo; Ströker, Erwin; Coutiño, Hugo-Enrique; Takarada, Ken; Salghetti, Francesca; Lusoc, Ian; Capulzini, Lucio; Brugada, Pedro; de Asmundis, Carlo; Chierchia, Gian-Battista

    2017-11-02

    The single-freeze strategy using the second-generation cryoballoon (CB-A, Arctic Front Advance, Medtronic, Minneapolis, MN, USA) has been reported to be as effective as the recommended double-freeze approach in several single-centre studies. In this retrospective, international, multicentre study, we compare the 3-min single-freeze strategy with the 4-min single-freeze strategy. Four hundred and thirty-two patients having undergone pulmonary vein isolation (PVI) by means of CB-A using a single-freeze strategy were considered for this analysis. A cohort of patients who were treated with a 3-min strategy (Group 1) was compared with a propensity score-matched cohort of patients who underwent a 4-min strategy (Group 2). Pulmonary vein isolation was successfully achieved in all the veins using the 28-mm CB-A. The procedural and fluoroscopy times were lower in Group 1 (67.8 ± 17 vs. 73.8 ± 26.3, P < 0.05; 14.9 ± 7.8 vs. 24.2 ± 10.6 min, P < 0.05). The most frequent complication was PNP, with no difference between the two groups (P = 0.67). After a mean follow-up of 13 ± 8 months, taking into consideration a blanking period of 3 months, 85.6% of patients in Group 1 and 87% of patients in Group 2 were free from arrhythmia recurrence at final follow-up (P = 0.67). There is no difference in acute success, rate of complications, and freedom from atrial fibrillation recurrences during the follow-up between 3-min and 4-min per vein freeze strategies. The procedural and fluoroscopy times were significantly shorter in 3-min per vein strategy. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2017. For Permissions, please email: journals.permissions@oup.com.

  7. Endolymphatic Thoracic Duct Stent-Graft Reconstruction for Chylothorax: Approach, Technical Success, Safety, and Short-term Outcomes.

    PubMed

    Srinivasa, Rajiv N; Chick, Jeffrey Forris Beecham; Hage, Anthony N; Gemmete, Joseph J; Murrey, Douglas C; Srinivasa, Ravi N

    2018-04-01

    To report approach, technical success, safety, and short-term outcomes of thoracic duct stent-graft reconstruction for the treatment of chylothorax. Two patients, 1 (50%) male and 1 (50%) female, with mean age of 38 years (range: 16-59 years) underwent endolymphatic thoracic duct stent-graft reconstruction between September 2016 and July 2017. Patients had radiographic left-sided chylothoraces (n = 2) from idiopathic causes (n = 1) and heart transplantation (n = 1). In both (100%) patients, antegrade lymphatic access was used to opacify the thoracic duct after which retrograde access was used for thoracic duct stent-graft placement. Pelvic lymphangiography technical success, antegrade cisterna chyli cannulation technical success, thoracic duct opacification technical success, retrograde thoracic duct access technical success, thoracic duct stent-graft reconstruction technical success, ethiodized oil volume, contrast volume, estimated blood loss, procedure time, fluoroscopy time, radiation dose, clinical success, complications, deaths, and follow-up were recorded. Pelvic lymphangiography, antegrade cisterna chyli cannulation, thoracic duct opacification, retrograde thoracic duct access, and thoracic duct stent-graft reconstruction were technically successful in both (100%) patients. Mean ethiodized oil volume was 8 mL (range: 5-10 mL). Mean contrast volume was 13 mL (range: 5-20 mL). Mean estimated blood loss was 13 mL (range: 10-15 mL). Mean fluoroscopy time was 50.4 min (range: 31.2-69.7 min). Mean dose area product and reference air kerma were 954.4 μGmy 2 (range: 701-1,208 μGmy 2 ) and 83.5 mGy (range: 59-108 mGy), respectively. Chylothorax resolved in both (100%) patients. There were no minor or major complications directly related to the procedure. Thoracic duct stent-graft reconstruction may be a technically successful and safe alternative to thoracic duct embolization, disruption, and surgical ligation for the treatment of chylothorax. Additional studies are warranted. Copyright © 2017 Elsevier Inc. All rights reserved.

  8. CBCT-based 3D MRA and angiographic image fusion and MRA image navigation for neuro interventions.

    PubMed

    Zhang, Qiang; Zhang, Zhiqiang; Yang, Jiakang; Sun, Qi; Luo, Yongchun; Shan, Tonghui; Zhang, Hao; Han, Jingfeng; Liang, Chunyang; Pan, Wenlong; Gu, Chuanqi; Mao, Gengsheng; Xu, Ruxiang

    2016-08-01

    Digital subtracted angiography (DSA) remains the gold standard for diagnosis of cerebral vascular diseases and provides intraprocedural guidance. This practice involves extensive usage of x-ray and iodinated contrast medium, which can induce side effects. In this study, we examined the accuracy of 3-dimensional (3D) registration of magnetic resonance angiography (MRA) and DSA imaging for cerebral vessels, and tested the feasibility of using preprocedural MRA for real-time guidance during endovascular procedures.Twenty-three patients with suspected intracranial arterial lesions were enrolled. The contrast medium-enhanced 3D DSA of target vessels were acquired in 19 patients during endovascular procedures, and the images were registered with preprocedural MRA for fusion accuracy evaluation. Low-dose noncontrasted 3D angiography of the skull was performed in the other 4 patients, and registered with the MRA. The MRA was overlaid afterwards with 2D live fluoroscopy to guide endovascular procedures.The 3D registration of the MRA and angiography demonstrated a high accuracy for vessel lesion visualization in all 19 patients examined. Moreover, MRA of the intracranial vessels, registered to the noncontrasted 3D angiography in the 4 patients, provided real-time 3D roadmap to successfully guide the endovascular procedures. Radiation dose to patients and contrast medium usage were shown to be significantly reduced.Three-dimensional MRA and angiography fusion can accurately generate cerebral vasculature images to guide endovascular procedures. The use of the fusion technology could enhance clinical workflow while minimizing contrast medium usage and radiation dose, and hence lowering procedure risks and increasing treatment safety.

  9. CBCT-based 3D MRA and angiographic image fusion and MRA image navigation for neuro interventions

    PubMed Central

    Zhang, Qiang; Zhang, Zhiqiang; Yang, Jiakang; Sun, Qi; Luo, Yongchun; Shan, Tonghui; Zhang, Hao; Han, Jingfeng; Liang, Chunyang; Pan, Wenlong; Gu, Chuanqi; Mao, Gengsheng; Xu, Ruxiang

    2016-01-01

    Abstract Digital subtracted angiography (DSA) remains the gold standard for diagnosis of cerebral vascular diseases and provides intraprocedural guidance. This practice involves extensive usage of x-ray and iodinated contrast medium, which can induce side effects. In this study, we examined the accuracy of 3-dimensional (3D) registration of magnetic resonance angiography (MRA) and DSA imaging for cerebral vessels, and tested the feasibility of using preprocedural MRA for real-time guidance during endovascular procedures. Twenty-three patients with suspected intracranial arterial lesions were enrolled. The contrast medium-enhanced 3D DSA of target vessels were acquired in 19 patients during endovascular procedures, and the images were registered with preprocedural MRA for fusion accuracy evaluation. Low-dose noncontrasted 3D angiography of the skull was performed in the other 4 patients, and registered with the MRA. The MRA was overlaid afterwards with 2D live fluoroscopy to guide endovascular procedures. The 3D registration of the MRA and angiography demonstrated a high accuracy for vessel lesion visualization in all 19 patients examined. Moreover, MRA of the intracranial vessels, registered to the noncontrasted 3D angiography in the 4 patients, provided real-time 3D roadmap to successfully guide the endovascular procedures. Radiation dose to patients and contrast medium usage were shown to be significantly reduced. Three-dimensional MRA and angiography fusion can accurately generate cerebral vasculature images to guide endovascular procedures. The use of the fusion technology could enhance clinical workflow while minimizing contrast medium usage and radiation dose, and hence lowering procedure risks and increasing treatment safety. PMID:27512846

  10. Safe and rapid isolation of pulmonary veins using a novel circular ablation catheter and duty-cycled RF generator.

    PubMed

    Fredersdorf, Sabine; Weber, Stefan; Jilek, Clemens; Heinicke, Norbert; VON Bary, Christian; Jungbauer, Carsten; Riegger, Günter A; Hamer, Okka W; Jeron, Andreas

    2009-10-01

    Ablation of atrial fibrillation (AF) has been one of the most difficult and time-consuming electrophysiological procedures. Due to the rapidly increasing demand for ablation procedures, technical advances would be helpful to reduce complexity and procedure time in AF ablation. Therefore, we investigated the feasibility of a single-catheter technique for pulmonary vein (PV) isolation utilizing a decapolar catheter combined with a duty-cycled, unipolar-bipolar radiofrequency (RF) generator. AF mapping and ablation was performed in 21 consecutive patients (mean age 59 +/- 12 years, 9 males) with paroxysmal AF (n = 17) and persistent AF (n = 4). The ablation catheter was forwarded to the LA via single-transseptal puncture. All electrodes were energized in 2 to 5 applications per vein, followed by segmental RF applications, as needed, to achieve electrical isolation. To assess left atrial anatomy for purposes of catheter manipulation, and later evaluate the possibility of asymptomatic PV-stenosis, CT or MR imaging was performed both prior to ablation and at 6-month follow-up. Isolation could be achieved in 85/86 veins (99%). Procedure time for ablation was 81 +/- 13 minutes, and fluoroscopy time was 30 +/- 11 minutes. There were no procedural complications. Success rate at 6 months was 86% (18/21). MR or CT imaging excluded asymptomatic PV-stenosis. Mapping and ablation of PVs can be performed in a safe and efficient manner using a single-catheter technique, with short procedure times and minimal learning curve. Thus, this system may be of high interest not only for high volume but all centers performing AF ablation.

  11. Radiation exposure of the anesthesiologist in the neurointerventional suite.

    PubMed

    Anastasian, Zirka H; Strozyk, Dorothea; Meyers, Philip M; Wang, Shuang; Berman, Mitchell F

    2011-03-01

    Scatter radiation during interventional radiology procedures can produce cataracts in participating medical personnel. Standard safety equipment for the radiologist includes eye protection. The typical configuration of fluoroscopy equipment directs radiation scatter away from the radiologist and toward the anesthesiologist. This study analyzed facial radiation exposure of the anesthesiologist during interventional neuroradiology procedures. Radiation exposure to the forehead of the anesthesiologist and radiologist was measured during 31 adult neuroradiologic procedures involving the head or neck. Variables hypothesized to affect anesthesiologist exposure were recorded for each procedure. These included total radiation emitted by fluoroscopic equipment, radiologist exposure, number of pharmacologic interventions performed by the anesthesiologist, and other variables. Radiation exposure to the anesthesiologist's face averaged 6.5 ± 5.4 μSv per interventional procedure. This exposure was more than 6-fold greater (P < 0.0005) than for noninterventional angiographic procedures (1.0 ± 1.0) and averaged more than 3-fold the exposure of the radiologist (ratio, 3.2; 95% CI, 1.8-4.5). Multiple linear regression analysis showed that the exposure of the anesthesiologist was correlated with the number of pharmacologic interventions performed by the anesthesiologist and the total exposure of the radiologist. Current guidelines for occupational radiation exposure to the eye are undergoing review and are likely to be lowered below the current 100-150 mSv/yr limit. Anesthesiologists who spend significant time in neurointerventional radiology suites may have ocular radiation exposure approaching that of a radiologist. To ensure parity with safety standards adopted by radiologists, these anesthesiologists should wear protective eyewear.

  12. Transcatheter closure of patent ductus arteriosus: 11 years of clinical experience in Cipto Mangunkusumo Hospital, Jakarta, Indonesia.

    PubMed

    Djer, Mulyadi M; Saputro, Dimas Dwi; Putra, Sukman Tulus; Idris, Nikmah Salamia

    2015-06-01

    Transcatheter closure of patent ductus arteriosus (PDA) has been suggested to be the standard treatment of PDA. Although, in general, the procedure shows a high successful rate, outcomes may vary among pediatric cardiology centers. To evaluate the effectiveness of transcatheter closure of PDA in Cipto Mangunkusumo Hospital, Jakarta, Indonesia, this was a retrospective study on patients who underwent transcatheter closure of PDA in Cipto Mangunkusumo Hospital during the period of 2002-2013. Hospital registry was reviewed and data about patients' characteristics, PDA severity, procedure, and outcomes were retrieved. There were 298 patients, of whom 90 were males, who underwent transcatheter closure of PDA during the study period. Median age was 3.4 years (1 months-18 years), and median body weight was 12 (3.6-59) kg. The diameter of PDA ranged from 1.1 to 15.4 mm with a median of 3.7 mm. Device could be deployed in all patients, in which most were the Amplatzer ductal occluder (69.8 %) and the remainders were coils. Median fluoroscopy time was 15.4 (1.5-87) min, and procedure time was 76 (30-200) min. Complete closure was achieved in most patients (97.3 %), whereas device migration occurred in a minority (0.3 %) of patients. No major complication occurred during or after the procedure. Transient anemia and bradycardia were found in 3.7 and 1.3 % patients, respectively. Most patients were discharged from the hospital at 1 day after the procedure. Transcatheter closure method is a safe and effective procedure to close PDA.

  13. Minimally invasive percutaneous nephrolithotomy: an effective treatment for kidney stones in infants under 1 year of age. A single-center experience.

    PubMed

    Dağgülli, Mansur; Sancaktutar, Ahmet Ali; Dede, Onur; Utanğaç, Mehmet Mazhar; Bodakçi, Mehmet Nuri; Penbegül, Necmettin; Hatipoğlu, Namık Kemal; Çakmakçı, Süleyman

    2015-11-01

    We aimed to present the outcomes of PNL surgery performed in infantile patients with small renal stones who were younger than 1 year of age. A single-center prospective trial was initiated and during the period between Jan 2013 and Jan 2015, PNL was applied to 20 renal units of 16 infants (6 girls and 10 boys), including 4 patients with bilateral kidney stones. PNL was performed in patients with renal stones larger than 2 cm, as well as stones resistant to SWL or renal stones that were undetectable during SWL. The mean age of the patients was 9.55 (5-12) months. Of the 20 renal units, 1 had complete staghorn stones, 3 had partial staghorn stones, 13 had renal pelvic stones, and 3 had lower pole stones. The mean stone size was 18.5 mm (range 12-36 mm). Mean operative time for PNL was 88 (25-135 min). Mean fluoroscopy time was estimated as 3.4 min. Mean hemoglobin loss was 0.72 g/L (0.2-3). The mean hospital stay was 4.1 days (2-8 days). On postoperative day 1, a complete stone-free state was achieved in 70% of renal units (14 of 20). At the end of the first postoperative week, the remaining two patients had insignificant residual fragments of 3 mm and were followed conservatively without any specific intervention. Thus, the total SFR was 80% (16 of 20) at discharge. In infants aged less than 1 year, minimal access tract dilation during PNL, the use of smaller caliber pediatric instruments, and the realization of this procedure by surgeons with adequate experience in adults carry utmost importance. In addition, special care should be taken to avoid hypothermia and radiation exposure during PNL.

  14. Optimization of block-floating-point realizations for digital controllers with finite-word-length considerations.

    PubMed

    Wu, Jun; Hu, Xie-he; Chen, Sheng; Chu, Jian

    2003-01-01

    The closed-loop stability issue of finite-precision realizations was investigated for digital controllers implemented in block-floating-point format. The controller coefficient perturbation was analyzed resulting from using finite word length (FWL) block-floating-point representation scheme. A block-floating-point FWL closed-loop stability measure was derived which considers both the dynamic range and precision. To facilitate the design of optimal finite-precision controller realizations, a computationally tractable block-floating-point FWL closed-loop stability measure was then introduced and the method of computing the value of this measure for a given controller realization was developed. The optimal controller realization is defined as the solution that maximizes the corresponding measure, and a numerical optimization approach was adopted to solve the resulting optimal realization problem. A numerical example was used to illustrate the design procedure and to compare the optimal controller realization with the initial realization.

  15. Direct percutaneous transaortic approach for treatment of aortic pseudoaneurysms.

    PubMed

    Pirelli, Luigi; Kliger, Chad; Fontana, Gregory P; Ruiz, Carlos E

    2015-05-01

    Aortic pseudoaneurysms (APAs) can develop months or years after aortic and cardiac surgery. If not treated appropriately, APAs can lead to fatal complications and ultimately death. We describe a case of a 61-year old patient with a diagnosed large pseudoaneurysm 5 years after his aortic valve surgery, who was treated with a novel transcatheter direct transaortic approach. The patient had dilated cardiomyopathy with an APA adjacent to the lower sternal plate. An Amplatzer septal occlusion device followed by coils was delivered transcutaneously through the APA to close its neck and fill the false aneurysm, respectively. Triple fusion multimodality imaging was used to guide the placement of the occlusion devices. The merging of computed tomography (CT) and echocardiography with real-time fluoroscopy was fundamental in procedural planning and guidance. Post-procedural transoesophageal echocardiogram (TOE) and CT angiography showed complete exclusion of the APA. A direct transaortic approach is a valid option for closure of an APA if the surgical risk is prohibitive, and the use of triple fusion technology is an essential tool in the hands of interventionalists and surgeons for preoperative planning and conduction of these procedures. © The Author 2015. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

  16. Past and present work practices of European interventional cardiologists in the context of radiation protection of the eye lens-results of the EURALOC study.

    PubMed

    Domienik-Andrzejewska, Joanna; Ciraj-Bjelac, Olivera; Askounis, Panagiotis; Covens, Peter; Dragusin, Octavian; Jacob, Sophie; Farah, Jad; Gianicolo, Emilio; Padovani, Renato; Teles, Pedro; Widmark, Anders; Struelens, Lara

    2018-05-21

    This paper investigates over five decades of work practices in interventional cardiology, with an emphasis on radiation protection. The analysis is based on data from more than 400 cardiologists from various European countries recruited for a EURALOC study and collected in the period from 2014 to 2016. Information on the types of procedures performed and their annual mean number, fluoroscopy time, access site choice, x-ray units and radiation protection means used was collected using an occupational questionnaire. Based on the specific European data, changes in each parameter have been analysed over decades, while country-specific data analysis has allowed us to determine the differences in local practices. In particular, based on the collected data, the typical workload of a European cardiologist working in a haemodynamic room and an electrophysiology room was specified for various types of procedures. The results showed that when working in a haemodynamic room, a transparent ceiling-suspended lead shield or lead glasses are necessary in order to remain below the recommended eye lens dose limit of 20 mSv. Moreover, the analysis revealed that new, more complex cardiac procedures such as chronic total occlusion, valvuloplasty and pulmonary vein isolation for atrial fibrillation ablation might contribute substantially to annual doses, although they are relatively rarely performed. The results revealed that considerable progress has been made in the use of radiation protection tools. While their use in electrophysiology procedures is not generic, the situation in haemodynamic procedures is rather encouraging, as ceiling-suspended shields are used in 90% of cases, while the combination of ceiling shield and lead glasses is noted in more than 40% of the procedures. However, we find that still 7% of haemodynamic procedures are performed without any radiation protection tools.

  17. Comparison of the feasibility and effectiveness of transradial coronary angiography via right versus left radial artery approaches (from the PREVAIL Study).

    PubMed

    Pelliccia, Francesco; Trani, Carlo; Biondi-Zoccai, Giuseppe G L; Nazzaro, Marco; Berni, Andrea; Patti, Giuseppe; Patrizi, Roberto; Pironi, Bruno; Mazzarotto, Pietro; Gioffrè, Gaetano; Speciale, Giulio; Pristipino, Christian

    2012-09-15

    It remains undefined if transradial coronary angiography from a right or left radial arterial approach differs in real-world practice. To address this issue, we performed a subanalysis of the PREVAIL study. The PREVAIL study was a prospective, multicenter, observational survey of unselected consecutive patients undergoing invasive cardiovascular procedures over a 1-month observation period, specifically aimed at assessing the outcomes of radial approach in the contemporary real world. The choice of arterial approach was left to the discretion of the operator. Prespecified end points of this subanalysis were procedural characteristics. Of 1,052 patients consecutively enrolled, 509 patients underwent transradial catheterization, 304 with a right radial and 205 with a left radial approach. Procedural success rates were similar between the 2 groups. Compared to the left radial group, the right radial group had longer procedure duration (46 ± 29 vs 33 ± 24 minutes, p <0.0001) and fluoroscopy time (765 ± 787 vs 533 ± 502, p <0.0001). At multivariate analysis, including a parsimonious propensity score for the choice of left radial approach, duration of procedure (beta coefficient 11.38, p <0.001) and total dose-area product (beta coefficient 11.38, p <0.001) were independently associated with the choice of the left radial artery approach. The operator's proficiency in right/left radial approach did not influence study results. In conclusion, right and left radial approaches are feasible and effective to perform percutaneous procedures. In the contemporary real world, however, the left radial route is associated with shorter procedures and lower radiologic exposure than the right radial approach, independently of an operator's proficiency. Copyright © 2012 Elsevier Inc. All rights reserved.

  18. Sacroplasty procedural extravasation with high viscosity bone cement: comparing the intraoperative long-axis versus short-axis techniques in osteoporotic cadavers.

    PubMed

    Miller, Jeffrey W; Diani, Art; Docsa, Steve; Ashton, Kristi; Sciamanna, Michele

    2017-09-01

    Percutaneous sacroplasty involves image-guided injection of bone cement for sacral insufficiency fractures to alleviate pain and facilitate mobility. Correct sacral placement of the cement and the risk of cement extravasation present procedural challenges. This study compares the occurrence, number, location, and surface area of high viscosity radiopaque bone cement extravasation via biplane fluoroscopy with Dyna CT between the fluoroscopically-guided intraoperative long-axis and short-axis sacroplasty techniques in osteoporotic cadavers. Ten osteoporotic cadavers underwent bilateral percutaneous instillation of VertaPlex HV High Viscosity Radiopaque Bone Cement. Long- and short-axis sacroplasty techniques were randomly assigned to zone 1 of the left or right sacral ala of each cadaver. Cement extravasation data were summarized by technique (long-axis vs short-axis) and time period (15-min and 3-hour post-procedure syngo DynaCT scan) in the form of point and CI estimates for the true proportions of cement extravasation. No procedural sacral extravasation differences were observed between the long-axis and short-axis sacroplasty techniques. There were no occurrences of intra-procedural or post-procedural cement extravasation at 15 min or 3 hours in association with either the long-axis sacroplasty technique or the short-axis sacroplasty technique. The long- and short-axis sacroplasty techniques, using high viscosity cement with careful post-procedural positioning, result in no occurrence of cement extravasation in porous osteoporotic cadaver bone. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/.

  19. Mini-STAR as bail-out strategy for percutaneous coronary intervention of chronic total occlusion.

    PubMed

    Galassi, Alfredo Ruggero; Tomasello, Salvatore Davide; Costanzo, Luca; Campisano, Maria Barbara; Barrano, Giombattista; Ueno, Masafumi; Tello-Montoliu, Antonio; Tamburino, Corrado

    2012-01-01

    Although the advancement of the equipment and the presence of innovative techniques, percutaneous coronary intervention (PCI) for chronic total occlusion (CTO) continues to be affected by lower procedural success in comparison with non occluded vessel PCI. We describe a new technique for the treatment of coronary CTO which utilizes a new generation of polymeric wires. From March 2009 to June 2010 different strategies were adopted as "bail out" after an initial attempt failed in 117 consecutive CTO lesions. Among these, conventional strategies (CS) such as parallel wire, sub-intimal tracking and re-entry (STAR), microchannel technique, intracoronary ultrasound guided revascularization and anchor balloon, were used in 75 cases (64.1%), while in the remaining a new technique, the "mini-STAR," was used (39.9%). Although no substantial differences were observed regarding the distribution of clinical features and angiographic lesions characteristics between the populations, mini-STAR was able to achieve a higher rate of procedural success in comparison with other CS (97.6% vs. 52%, P < 0.001) with lower contrast agent use (442 ± 259 cm(3) vs. 561 ± 243 cm(3), P = 0.01) and shorter procedural and fluoroscopy times (122 ± 61 vs. 157 ± 74 min, P = 0.009 and 60 ± 31 min vs. 75 ± 38 min, P = 0.03, respectively). No differences were observed in term of peri-procedural complications such as procedural myocardial infarction, coronary perforations, and contrast-induced nephropathy between mini-STAR and CS. The mini-STAR technique is a promising strategy for the treatment of CTO lesions, achieving a high procedural success rate and low occurrence of procedural adverse events. Copyright © 2011 Wiley Periodicals, Inc.

  20. Directly ventricular septal defect closure without using arteriovenous wire loop: Our adult case series using transarterial retrograde approach

    PubMed Central

    Pekel, Nihat; Ercan, Ertuğrul; Özpelit, Mehmet Emre; Özyurtlu, Ferhat; Yılmaz, Akar; Topaloğlu, Caner; Saygı, Serkan; Yakan, Serkan; Tengiz, İstemihan

    2017-01-01

    Objective: The standard transcatheter ventricular septal defects (VSD) closure procedure is established with arteriovenous (AV) loop and is called as antegrade approach. The directly retrograde transarterial VSD closure without using AV loop might be better option as shortens the procedure time and decreases radiation exposure. Methods: Our series consist of twelve sequential adult cases with congenital VSDs (seven with perimembranous, four with muscular, one with postoperative residuel VSD). The mean age was 26.9 (Range 18–58), the mean height was 168.75 cm (Range 155–185cm), and the mean body mass index was 23.4 (Range 17.3–28.4). Maximum and minimum defect sizes were 10 and 5 mm and the mean defect size was 6.24 mm. The procedure was performed with left heart catheterization and advancing the delivery sheath over the stiff exchange wire then VSD occlusion from left side. Results: The defects were successfully closed with this technique in eleven patients. In sixth patient, the defect could not be cannulated by the delivery sheath, as the tip of the sheath did not reach the defect and VSD was closed with same sheath by standard transvenous approach using AV loop. We didn’t encounter any complication releated to semilunar or atrioventricular valves. Atrioventricular conduction system was not affected by the procedure in any patients. The median procedure and fluoroscopy times were 66 and 16.5 minutes respectively. Conclusion: Transarterial retrograde VSD closure without using AV loop simplifies the procedure, decreases the radiation exposure, and shortens the procedure time. The only limitation in adult patients is delivery sheath length. PMID:28315566

Top