Science.gov

Sample records for ionization chamber smoke detectors

  1. Ionization-chamber smoke detector system

    DOEpatents

    Roe, Robert F.

    1976-10-19

    This invention relates to an improved smoke-detection system of the ionization-chamber type. In the preferred embodiment, the system utilizes a conventional detector head comprising a measuring ionization chamber, a reference ionization chamber, and a normally non-conductive gas triode for discharging when a threshold concentration of airborne particulates is present in the measuring chamber. The improved system is designed to reduce false alarms caused by fluctuations in ambient temperature. Means are provided for periodically firing the gas discharge triode and each time recording the triggering voltage required. A computer compares each triggering voltage with its predecessor. The computer is programmed to energize an alarm if the difference between the two compared voltages is a relatively large value indicative of particulates in the measuring chamber and to disregard smaller differences typically resulting from changes in ambient temperature.

  2. IONIZATION CHAMBER

    DOEpatents

    Redman, W.C.; Shonka, F.R.

    1958-02-18

    This patent describes a novel ionization chamber which is well suited to measuring the radioactivity of the various portions of a wire as the wire is moved at a uniform speed, in order to produce the neutron flux traverse pattern of a reactor in which the wire was previously exposed to neutron radiation. The ionization chamber of the present invention is characterized by the construction wherein the wire is passed through a tubular, straight electrode and radiation shielding material is disposed along the wire except at an intermediate, narrow area where the second electrode of the chamber is located.

  3. Ionization chamber

    DOEpatents

    Walenta, Albert H.

    1981-01-01

    An ionization chamber has separate drift and detection regions electrically isolated from each other by a fine wire grid. A relatively weak electric field can be maintained in the drift region when the grid and another electrode in the chamber are connected to a high voltage source. A much stronger electric field can be provided in the detection region by connecting wire electrodes therein to another high voltage source. The detection region can thus be operated in a proportional mode when a suitable gas is contained in the chamber. High resolution output pulse waveforms are provided across a resistor connected to the detection region anode, after ionizing radiation enters the drift region and ionize the gas.

  4. On line high dose static position monitoring by ionization chamber detector for industrial gamma irradiators.

    PubMed

    Rodrigues, Ary A; Vieira, Jose M; Hamada, Margarida M

    2010-01-01

    A 1 cm(3) cylindrical ionization chamber was developed to measure high doses on line during the sample irradiation in static position, in a (60)Co industrial plant. The developed ionization chamber showed to be suitable for use as a dosimeter on line. A good linearity of the detector was found between the dose and the accumulated charge, independently of the different dose rates caused by absorbing materials.

  5. Neutron and gamma detector using an ionization chamber with an integrated body and moderator

    DOEpatents

    Ianakiev, Kiril D.; Swinhoe, Martyn T.; Lestone, John Paul

    2006-07-18

    A detector for detecting neutrons and gamma radiation includes a cathode that defines an interior surface and an interior volume. A conductive neutron-capturing layer is disposed on the interior surface of the cathode and a plastic housing surrounds the cathode. A plastic lid is attached to the housing and encloses the interior volume of the cathode forming an ionization chamber, into the center of which an anode extends from the plastic lid. A working gas is disposed within the ionization chamber and a high biasing voltage is connected to the cathode. Processing electronics are coupled to the anode and process current pulses which are converted into Gaussian pulses, which are either counted as neutrons or integrated as gammas, in response to whether pulse amplitude crosses a neutron threshold. The detector according to the invention may be readily fabricated into single or multilayer detector arrays.

  6. Dosimetric characteristics of the novel 2D ionization chamber array OCTAVIUS Detector 1500

    SciTech Connect

    Stelljes, T. S. Looe, H. K.; Chofor, N.; Poppe, B.; Harmeyer, A.; Reuter, J.; Harder, D.

    2015-04-15

    Purpose: The dosimetric properties of the OCTAVIUS Detector 1500 (OD1500) ionization chamber array (PTW-Freiburg, Freiburg, Germany) have been investigated. A comparative study was carried out with the OCTAVIUS Detector 729 and OCTAVIUS Detector 1000 SRS arrays. Methods: The OD1500 array is an air vented ionization chamber array with 1405 detectors in a 27 × 27 cm{sup 2} measurement area arranged in a checkerboard pattern with a chamber-to-chamber distance of 10 mm in each row. A sampling step width of 5 mm can be achieved by merging two measurements shifted by 5 mm, thus fulfilling the Nyquist theorem for intensity modulated dose distributions. The stability, linearity, and dose per pulse dependence were investigated using a Semiflex 31013 chamber (PTW-Freiburg, Freiburg, Germany) as a reference detector. The effective depth of measurement was determined by measuring TPR curves with the array and a Roos chamber type 31004 (PTW-Freiburg, Freiburg, Germany). Comparative output factor measurements were performed with the array, the Semiflex 31010 ionization chamber and the Diode 60012 (both PTW-Freiburg, Freiburg, Germany). The energy dependence of the OD1500 was measured by comparing the array’s readings to those of a Semiflex 31010 ionization chamber for varying mean photon energies at the depth of measurement, applying to the Semiflex chamber readings the correction factor k{sub NR} for nonreference conditions. The Gaussian lateral dose response function of a single array detector was determined by searching the convolution kernel suitable to convert the slit beam profiles measured with a Diode 60012 into those measured with the array’s central chamber. An intensity modulated dose distribution measured with the array was verified by comparing a OD1500 measurement to TPS calculations and film measurements. Results: The stability and interchamber sensitivity variation of the OD1500 array were within ±0.2% and ±0.58%, respectively. Dose linearity was within 1

  7. An ionization chamber and a Si-detector for lead-210 chronology

    NASA Astrophysics Data System (ADS)

    Farid, M.; El-Daoushy, A. F.; El-Daoushy, M. F. A. F.

    1981-09-01

    Radon emanation and isotopic dilution techniques were used for the determination of 226Ra and "total" 210Pb in sediment samples. The "unsupported" 210Pb were then used to construct lake-sediment chronologies. Polonium was extracted at 550-600°C, transferred to chloride, then plated by self-deposition on silver disks. Memory effects due to adsorption of polonium on glass were carefully studied. Si-detectors were used for α-activity measurements. Background studies indicated instabilities in a Si-detector when left without bias for longer periods. The radium was extracted with the help of a barium carrier, which not only made the extraction quantitative but also eliminated the adsorption of radium on the glasses used. The 222Rn was measured in an ionization chamber. The gas counter with its modified filling system allowed both low level measurements of sediment samples and counter calibration with comparably active 226Ra standards. The increasing background of the ionization chamber was explained by the adsorption of 222Rn on the surfaces of the counter. The background was reduced by the removal of the adsorbed atoms. Normalization for discrimination shifts, due to electronegative impurities, is required in case the counting gas is impure.

  8. Evaluation of smoke detectors for mining use. Report of investigations/1995

    SciTech Connect

    Edwards, J.C.; Morrow, G.S.

    1995-09-01

    The U.S. Bureau of Mines has constructed a smoke chamber and developed sensitivity tests for smoke detectors. Response of ionization- and optical-type commercially available smoke detectors have been investigated. Six smoke detectors were measured with respect to visually obscuring smoke characterized by a corresponding optical density for smoldering and flaming coal combustion in the smoke chamber. The responsive characteristics of the detectors evaluated with respect to known smoke conditions in the smoke chamber shows their potential for use as mine fire sensors or part of a mine atmospheric monitoring system to improve mine safety.

  9. The magic cube and the pixel ionization chamber: detectors for monitor and dosimetry of radiotherapy beams

    NASA Astrophysics Data System (ADS)

    Amerio, S.; Boriano, A.; Bourhaleb, F.; Cirio, R.; Donetti, M.; Garelli, E.; Giordanengo, S.; Madon, E.; Marchetto, F.; Nastasi, U.; Peroni, C.; Sanz Freire, C. J.; Sardo, A.; Trevisiol, E.

    2003-09-01

    Tumor therapy takes advantage of the energy deposition of radiation to concentrate high doses in the target while sparing healthy tissue. Elective pathologies for highly conformal radiotherapies such as photon Intensity Modulated Radiotherapy (IMRT) and radiotherapy with hadrons are head and neck, eye, prostate and in general all tumors that are either deep or located close to critical organs. In the world there are several centers that are using such techniques and a common problem that is being experienced is the verification of treatment plans and monitoring of the beam. We have designed and built two detectors that allow 2D and 3D measurements of dose and fluence of such beams. The detectors allow measurements on big surfaces, up to 25∗25 cm2. The active media are parallel plate, strip and pixel segmented ionization chambers with front-end Very Large Scale Integration (VLSI) readout and PC based data acquistion. The description of dosimeter, chamber and electronics will be given with results from beam tests and therapy plan verification.

  10. Smoke Detector

    NASA Technical Reports Server (NTRS)

    1979-01-01

    In the photo, Fire Chief Jay Stout of Safety Harbor, Florida, is explaining to young Richard Davis the workings of the Honeywell smoke and fire detector which probably saved Richard's life and that of his teen-age brother. Alerted by the detector's warning, the pair were able to escape their burning home. The detector in the Davis home was one of 1,500 installed in Safety Harbor residences in a cooperative program conducted by the city and Honeywell Inc.

  11. Development of coal combustion sensitivity test for smoke detectors

    SciTech Connect

    Edwards, J.C.; Morrow, G.S.

    1995-09-01

    Standard smoldering and flaming combustion tests using small coal samples have been developed by the US Bureau of Mines as a method to evaluate the response of a smoke detector. The tests are conducted using a standard smoke box designed and constructed according to Underwriters Laboratories. The tests provide a standard, easily reproducible smoke characteristic for smoldering and flaming coal combustion, based upon a comparison of the smoke optical density and the response of a standard ionization chamber to the smoke. With these standard tests, the range of threshold limits for the response of a smoke detector and the detector`s reliability can be evaluated for nearly identical smoke visibility and smoke physical characteristics. The detector`s threshold response limits and reliability need to be well defined prior to the instrument`s use as part of a mine fire warning system for improved mine safety.

  12. A combined segmented anode gas ionization chamber and time-of-flight detector for heavy ion elastic recoil detection analysis

    NASA Astrophysics Data System (ADS)

    Ström, Petter; Petersson, Per; Rubel, Marek; Possnert, Göran

    2016-10-01

    A dedicated detector system for heavy ion elastic recoil detection analysis at the Tandem Laboratory of Uppsala University is presented. Benefits of combining a time-of-flight measurement with a segmented anode gas ionization chamber are demonstrated. The capability of ion species identification is improved with the present system, compared to that obtained when using a single solid state silicon detector for the full ion energy signal. The system enables separation of light elements, up to Neon, based on atomic number while signals from heavy elements such as molybdenum and tungsten are separated based on mass, to a sample depth on the order of 1 μm. The performance of the system is discussed and a selection of material analysis applications is given. Plasma-facing materials from fusion experiments, in particular metal mirrors, are used as a main example for the discussion. Marker experiments using nitrogen-15 or oxygen-18 are specific cases for which the described improved species separation and sensitivity are required. Resilience to radiation damage and significantly improved energy resolution for heavy elements at low energies are additional benefits of the gas ionization chamber over a solid state detector based system.

  13. Correction of measured Gamma-Knife output factors for angular dependence of diode detectors and PinPoint ionization chamber.

    PubMed

    Hršak, Hrvoje; Majer, Marija; Grego, Timor; Bibić, Juraj; Heinrich, Zdravko

    2014-12-01

    Dosimetry for Gamma-Knife requires detectors with high spatial resolution and minimal angular dependence of response. Angular dependence and end effect time for p-type silicon detectors (PTW Diode P and Diode E) and PTW PinPoint ionization chamber were measured with Gamma-Knife beams. Weighted angular dependence correction factors were calculated for each detector. The Gamma-Knife output factors were corrected for angular dependence and end effect time. For Gamma-Knife beams angle range of 84°-54°. Diode P shows considerable angular dependence of 9% and 8% for the 18 mm and 14, 8, 4 mm collimator, respectively. For Diode E this dependence is about 4% for all collimators. PinPoint ionization chamber shows angular dependence of less than 3% for 18, 14 and 8 mm helmet and 10% for 4 mm collimator due to volumetric averaging effect in a small photon beam. Corrected output factors for 14 mm helmet are in very good agreement (within ±0.3%) with published data and values recommended by vendor (Elekta AB, Stockholm, Sweden). For the 8 mm collimator diodes are still in good agreement with recommended values (within ±0.6%), while PinPoint gives 3% less value. For the 4 mm helmet Diodes P and E show over-response of 2.8% and 1.8%, respectively. For PinPoint chamber output factor of 4 mm collimator is 25% lower than Elekta value which is generally not consequence of angular dependence, but of volumetric averaging effect and lack of lateral electronic equilibrium. Diodes P and E represent good choice for Gamma-Knife dosimetry.

  14. Large area neutron detector based on Li6 ionization chamber with integrated body-moderator of high density polyethylene

    SciTech Connect

    Ianakiev, Kiril D.; Swinhoe, Martyn T.; Chung, Kiwhan; Makela, Mark F.

    2009-06-30

    A detector was developed and funded by DHS to be a lower cost alternative to 3He detectors. A 6Li foil-lined ionization chamber was prepared with fill gas at one atmosphere and pulse mode operation. The high-density polyethylene (HOPE) body serves also as a neutron moderator. All electrodes, including high voltage bias supply, are hermetically sealed within the plastic slabs.

  15. Liquid-filled ionization chamber temperature dependence

    NASA Astrophysics Data System (ADS)

    Franco, L.; Gómez, F.; Iglesias, A.; Pardo, J.; Pazos, A.; Pena, J.; Zapata, M.

    2006-05-01

    Temperature and pressure corrections of the read-out signal of ionization chambers have a crucial importance in order to perform high-precision absolute dose measurements. In the present work the temperature and pressure dependences of a sealed liquid isooctane filled ionization chamber (previously developed by the authors) for radiotherapy applications have been studied. We have analyzed the thermal response of the liquid ionization chamber in a ˜20C interval around room temperature. The temperature dependence of the signal can be considered linear, with a slope that depends on the chamber collection electric field. For example, a relative signal slope of 0.27×10-2 K-1 for an operation electric field of 1.67×106 V m-1 has been measured in our detector. On the other hand, ambient pressure dependence has been found negligible, as expected for liquid-filled chambers. The thermal dependence of the liquid ionization chamber signal can be parametrized within the Onsager theory on initial recombination. Considering that changes with temperature of the detector response are due to variations in the free ion yield, a parametrization of this dependence has been obtained. There is a good agreement between the experimental data and the theoretical model from the Onsager framework.

  16. Comparison of depth-dose distributions of proton therapeutic beams calculated by means of logical detectors and ionization chamber modeled in Monte Carlo codes

    NASA Astrophysics Data System (ADS)

    Pietrzak, Robert; Konefał, Adam; Sokół, Maria; Orlef, Andrzej

    2016-08-01

    The success of proton therapy depends strongly on the precision of treatment planning. Dose distribution in biological tissue may be obtained from Monte Carlo simulations using various scientific codes making it possible to perform very accurate calculations. However, there are many factors affecting the accuracy of modeling. One of them is a structure of objects called bins registering a dose. In this work the influence of bin structure on the dose distributions was examined. The MCNPX code calculations of Bragg curve for the 60 MeV proton beam were done in two ways: using simple logical detectors being the volumes determined in water, and using a precise model of ionization chamber used in clinical dosimetry. The results of the simulations were verified experimentally in the water phantom with Marcus ionization chamber. The average local dose difference between the measured relative doses in the water phantom and those calculated by means of the logical detectors was 1.4% at first 25 mm, whereas in the full depth range this difference was 1.6% for the maximum uncertainty in the calculations less than 2.4% and for the maximum measuring error of 1%. In case of the relative doses calculated with the use of the ionization chamber model this average difference was somewhat greater, being 2.3% at depths up to 25 mm and 2.4% in the full range of depths for the maximum uncertainty in the calculations of 3%. In the dose calculations the ionization chamber model does not offer any additional advantages over the logical detectors. The results provided by both models are similar and in good agreement with the measurements, however, the logical detector approach is a more time-effective method.

  17. Alkali metal ionization detector

    DOEpatents

    Bauerle, James E.; Reed, William H.; Berkey, Edgar

    1978-01-01

    Variations in the conventional filament and collector electrodes of an alkali metal ionization detector, including the substitution of helical electrode configurations for either the conventional wire filament or flat plate collector; or, the substitution of a plurality of discrete filament electrodes providing an in situ capability for transferring from an operationally defective filament electrode to a previously unused filament electrode without removing the alkali metal ionization detector from the monitored environment. In particular, the helical collector arrangement which is coaxially disposed about the filament electrode, i.e. the thermal ionizer, provides an improved collection of positive ions developed by the filament electrode. The helical filament design, on the other hand, provides the advantage of an increased surface area for ionization of alkali metal-bearing species in a monitored gas environment as well as providing a relatively strong electric field for collecting the ions at the collector electrode about which the helical filament electrode is coaxially positioned. Alternatively, both the filament and collector electrodes can be helical. Furthermore, the operation of the conventional alkali metal ionization detector as a leak detector can be simplified as to cost and complexity, by operating the detector at a reduced collector potential while maintaining the sensitivity of the alkali metal ionization detector adequate for the relatively low concentration of alkali vapor and aerosol typically encountered in leak detection applications.

  18. Smoke and particle detector using tritiated semiconductor foil

    SciTech Connect

    Liu, B.; Alvarez-Ossa, D.; Kherani, N. P.; Zukotynski, S.; Chen, K. P.

    2008-07-15

    A smoke and particle ionization detector using tritiated amorphous silicon film as the radiation source is demonstrated. The ion chamber design includes both bipolar and unipolar region; the unipolar region is defined as the volume space extending beyond the range of ionizing particles. Attachment of ions to particulate matter in the unipolar region considerably reduces the mobility of the carriers, thus forming a space charge cloud accompanied by a reduction in the electrical field strength, thereby enhancing the particulate detection responsivity. Tritium {beta}s have a maximum range of about 6 mm in air, which makes the detector compact. Owing to the potential of increased specific activity, it results in a gamma-free detector with improved sensitivity. The results show that this gamma-free detector is several-fold to forty-fold more responsive than traditional ionization detectors using Am-241. In addition, this ion chamber can function as a dual detector having both photoelectric and ionization detector responsivities. (authors)

  19. Ionization chamber dosimeter

    DOEpatents

    Renner, Tim R.; Nyman, Mark A.; Stradtner, Ronald

    1991-01-01

    A method for fabricating an ion chamber dosimeter collecting array of the type utilizing plural discrete elements formed on a uniform collecting surface which includes forming a thin insulating layer over an aperture in a frame having surfaces, forming a predetermined pattern of through holes in the layer, plating both surfaces of the layer and simultaneously tilting and rotating the frame for uniform plate-through of the holes between surfaces. Aligned masking and patterned etching of the surfaces provides interconnects between the through holes and copper leads provided to external circuitry.

  20. Alkali ionization detector

    DOEpatents

    Hrizo, John; Bauerle, James E.; Witkowski, Robert E.

    1982-01-01

    A calibration filament containing a sodium-bearing compound is included in combination with the sensing filament and ion collector plate of a sodium ionization detector to permit periodic generation of sodium atoms for the in-situ calibration of the detector.

  1. Radiation damage to tetramethylsilane and tetramethylgermanium ionization chambers

    SciTech Connect

    Hoshi, Y.; Higuchi, M.; Oyama, K. . Dept. of Applied Physics)

    1994-08-01

    Two detector media suitable for a warm liquid, ionization chamber filled with tetramethylsilane (TMS) and tetramethylgermanium (TMG) were exposed to [gamma] radiation form a [sup 60]Co source up to dose 579 Gray and 902 Gray, respectively. The electron lifetimes and the free ion yields were measured as a function of accumulated radiation dose. A similar behavior of the electron lifetimes and the free ion yields with increasing radiation does was observed between the TMS and TMG ionization chambers.

  2. DETECTORS AND EXPERIMENTAL METHODS: Design and construction of the first prototype ionization chamber for CSNS and PA beam loss monitor (BLM) system

    NASA Astrophysics Data System (ADS)

    Xu, Mei-Hang; Tian, Jian-Min; Chen, Chang; Chen, Yuan-Bo; Xu, Tao-Guang; Lu, Shuang-Tong

    2009-02-01

    Design and construction of the first prototype ionization chamber for CSNS and Proton Accelerator (PA) beam loss monitor (BLM) system is reported. The low leakage current (<0.1 pA), good plateau (approx800 V) and linearity range up to 200 Roentgen/h are obtained in the first prototype. All of these give us good experience for further improving the ionization chamber construction.

  3. Ionizing radiation detector

    DOEpatents

    Thacker, Louis H.

    1990-01-01

    An ionizing radiation detector is provided which is based on the principle of analog electronic integration of radiation sensor currents in the sub-pico to nano ampere range between fixed voltage switching thresholds with automatic voltage reversal each time the appropriate threshold is reached. The thresholds are provided by a first NAND gate Schmitt trigger which is coupled with a second NAND gate Schmitt trigger operating in an alternate switching state from the first gate to turn either a visible or audible indicating device on and off in response to the gate switching rate which is indicative of the level of radiation being sensed. The detector can be configured as a small, personal radiation dosimeter which is simple to operate and responsive over a dynamic range of at least 0.01 to 1000 R/hr.

  4. Modulated voltage metastable ionization detector

    NASA Technical Reports Server (NTRS)

    Carle, G. C.; Kojiro, D. R.; Humphrey, D. E. (Inventor)

    1985-01-01

    The output current from a metastable ionization detector (MID) is applied to a modulation voltage circuit. An adjustment is made to balance out the background current, and an output current, above background, is applied to an input of a strip chart recorder. For low level concentrations, i.e., low detected output current, the ionization potential will be at a maximum and the metastable ionization detector will operate at its most sensitive level. When the detected current from the metastable ionization detector increases above a predetermined threshold level, a voltage control circuit is activated which turns on a high voltage transistor which acts to reduce the ionization potential. The ionization potential applied to the metastable ionization detector is then varied so as to maintain the detected signal level constant. The variation in ionization potential is now related to the concentration of the constituent and a representative amplitude is applied to another input of said strip chart recorder.

  5. Ion chamber based neutron detectors

    DOEpatents

    Derzon, Mark S; Galambos, Paul C; Renzi, Ronald F

    2014-12-16

    A neutron detector with monolithically integrated readout circuitry, including: a bonded semiconductor die; an ion chamber formed in the bonded semiconductor die; a first electrode and a second electrode formed in the ion chamber; a neutron absorbing material filling the ion chamber; and the readout circuitry which is electrically coupled to the first and second electrodes. The bonded semiconductor die includes an etched semiconductor substrate bonded to an active semiconductor substrate. The readout circuitry is formed in a portion of the active semiconductor substrate. The ion chamber has a substantially planar first surface on which the first electrode is formed and a substantially planar second surface, parallel to the first surface, on which the second electrode is formed. The distance between the first electrode and the second electrode may be equal to or less than the 50% attenuation length for neutrons in the neutron absorbing material filling the ion chamber.

  6. Optical ionization detector

    DOEpatents

    Wuest, Craig R.; Lowry, Mark E.

    1994-01-01

    An optical ionization detector wherein a beam of light is split so that one arm passes through a fiber optics and the other arm passes through a gas-filled region, and uses interferometry to detect density changes in a gas when charged particles pass through it. The gas-filled region of the detector is subjected to a high electric field and as a charged particle traverses this gas region electrons are freed from the cathode and accelerated so as to generate an electron avalanche which is collected on the anode. The gas density is effected by the electron avalanche formation and if the index or refraction is proportional to the gas density the index will change accordingly. The detector uses this index change by modulating the one arm of the split light beam passing through the gas, with respect to the other arm that is passed through the fiber optic. Upon recombining of the beams, interference fringe changes as a function of the index change indicates the passage of charged particles through the gaseous medium.

  7. Optical ionization detector

    DOEpatents

    Wuest, C.R.; Lowry, M.E.

    1994-03-29

    An optical ionization detector wherein a beam of light is split so that one arm passes through a fiber optics and the other arm passes through a gas-filled region, and uses interferometry to detect density changes in a gas when charged particles pass through it. The gas-filled region of the detector is subjected to a high electric field and as a charged particle traverses this gas region electrons are freed from the cathode and accelerated so as to generate an electron avalanche which is collected on the anode. The gas density is effected by the electron avalanche formation and if the index or refraction is proportional to the gas density the index will change accordingly. The detector uses this index change by modulating the one arm of the split light beam passing through the gas, with respect to the other arm that is passed through the fiber optic. Upon recombining of the beams, interference fringe changes as a function of the index change indicates the passage of charged particles through the gaseous medium. 3 figures.

  8. The effect of a compression paddle on energy response, calibration and measurement with mammographic dosimeters using ionization chambers and solid-state detectors.

    PubMed

    Hourdakis, C J; Boziari, A; Koumbouli, E

    2009-02-21

    A compression paddle is always used in mammography x-ray examinations, in order to improve image quality and reduce patient doses. Although clinical dose measurements should be performed with the paddle to interfere with the x-ray beam, calibration of mammography dosimeters is performed free in air without the presence of the paddle. The paddle hardens the x-ray beam, which has an impact on a dosimeter performance, particularly on high-energy-dependent detectors. Due to the paddle, clinical mammography x-ray systems may exhibit beams with HVL values exceeding those of the IEC 61267 RQR-M series qualities at which dosimeters are usually calibrated. In this study, the influence of the paddle in mammography dosimetry is examined, in Mo/Mo anode/filter x-ray qualities. PMMA slabs of 1, 2 and 3 mm thickness and Al foils of 0.05, 0.10 and 0.15 mm thicknesses were used to simulate the paddles, producing beams with HVL values from 0.28 up to 0.43 mmAl. In these qualities, four solid-state (ST) detectors and three ionizations chambers (IC) were calibrated in terms of Kair and N(K) and k(Q) were deduced. The results showed that all IC and two modern-type ST dosimeters have a flat energy response in the above HVL range (less than 3%), so their calibration factor at RQR-M2 quality could be safely used for clinical measurements. Two other ST dosimeters exhibit up to 20% energy response, so differences up to 15% in dose measurement may be observed if the effect of paddle on their performance is ignored. Finally, the need of additional mammographic calibration qualities to the existing IEC 61267 RQR-M series is examined and discussed.

  9. Segmented ionization chambers for beam monitoring in hadrontherapy

    NASA Astrophysics Data System (ADS)

    Braccini, Saverio; Cirio, Roberto; Donetti, Marco; Marchetto, Flavio; Pittà, Giuseppe; Lavagno, Marco; La Rosa, Vanessa

    2015-05-01

    Segmented ionization chambers represent a good solution to monitor the position, the intensity and the shape of ion beams in hadrontherapy. Pixel and strip chambers have been developed for both passive scattering and active scanning dose delivery systems. In particular, strip chambers are optimal for pencil beam scanning, allowing for spatial and time resolutions below 0.1 mm and 1 ms, respectively. The MATRIX pixel and the Strip Accurate Monitor for Beam Applications (SAMBA) detectors are described in this paper together with the results of several beam tests and industrial developments based on these prototypes.

  10. Ionization-based detectors for gas chromatography.

    PubMed

    Poole, Colin F

    2015-11-20

    The gas phase ionization detectors are the most widely used detectors for gas chromatography. The column and makeup gases commonly used in gas chromatography are near perfect insulators. This facilitates the detection of a minute number of charge carriers facilitating the use of ionization mechanisms of low efficiency while providing high sensitivity. The main ionization mechanism discussed in this report are combustion in a hydrogen diffusion flame (flame ionization detector), surface ionization in a plasma (thermionic ionization detector), photon ionization (photoionization detector and pulsed discharge helium ionization detector), attachment of thermal electrons (electron-capture detector), and ionization by collision with metastable helium species (helium ionization detector). The design, response characteristics, response mechanism, and suitability for fast gas chromatography are the main features summarized in this report. Mass spectrometric detection and atomic emission detection, which could be considered as ionization detectors of a more sophisticated and complex design, are not discussed in this report.

  11. Ionization Chamber for Prompt Fission Neutron Investigations

    NASA Astrophysics Data System (ADS)

    Zeynalov, Sh.; Zeynalova, O.; Hambsch, F.-J.; Sedyshev, P.; Shvetsov, V.

    In this work we report recent achievements in design of twin back-to-back ionization chamber (TIC) for fission fragment (FF) mass and kinetic energy measurement. Correlated FF kinetic energies, their masses and the angle of FF in respect to the axes in 3D Cartesian coordinates can be determined from analysis of the heights and shapes of the pulses induced by the fission fragments on the anodes of TIC. Anodes of TIC were designed as consisting of isolated strips each having independent electronic circuitry and special multi-channel pulse processing apparatus. Mathematical formulae provided for FF angles measured in respect to the coordinate axes. It was shown how the point of fission fragments origin on the target plane may be determined using the same measured data. The last feature made the TIC a rather powerful tool for prompt fission neutron (PFN) emission investigation in event-by-event analysis of individual fission reactions from non- point fissile source. Position sensitive neutron induced fission detector for neutron-imaging applications with both thermal and low energy neutrons was found as another possible implementation of the designed TIC.

  12. 24 CFR 200.76 - Smoke detectors.

    Code of Federal Regulations, 2010 CFR

    2010-04-01

    ... DEVELOPMENT GENERAL INTRODUCTION TO FHA PROGRAMS Requirements for Application, Commitment, and Endorsement... Eligibility Requirements for Existing Projects Property Requirements § 200.76 Smoke detectors. Smoke...

  13. 24 CFR 200.76 - Smoke detectors.

    Code of Federal Regulations, 2011 CFR

    2011-04-01

    ... DEVELOPMENT GENERAL INTRODUCTION TO FHA PROGRAMS Requirements for Application, Commitment, and Endorsement... Eligibility Requirements for Existing Projects Property Requirements § 200.76 Smoke detectors. Smoke...

  14. a Liquid Ionization Chamber as Monitor in Radiotherapy

    NASA Astrophysics Data System (ADS)

    Berghöfer, Th.; Engler, J.; Milke, J. M.; Hörandel, J. R.; Hartmann, G. H.

    2006-04-01

    First measurements with a prototype liquid ionization chamber are described to be applied as an online-monitor for intensity modulated radiotherapy. The detector consists of 480 individual electronic channels which allow parallel read-out of radiation induced currents at frequencies exceeding 10 Hz. Dose gradients in the direction of leaf movement of a multileaf collimator have been measured and a reconstruction method for individual leaf positions has been developed. The achieved reconstruction accuracy will be described.

  15. Neutron-chamber detectors and applications

    SciTech Connect

    Fehlau, P.E.; Atwater, H.F.; Coop, K.L.

    1990-01-01

    Detector applications in Nuclear Safeguards and Waste Management have included measuring neutrons from fission and (alpha,n) reactions with well-moderated neutron proportional counters, often embedded in a slab of polyethylene. Other less-moderated geometries are useful for detecting both bare and moderated fission-source neutrons with good efficiency. The neutron chamber is an undermoderated detector design comprising a large, hollow, polyethylene-walled chamber containing one or more proportional counters. Neutron-chamber detectors are relatively inexpensive; can have large apertures, usually through a thin chamber wall; and offer very good detection efficiency per dollar. Neutron-chamber detectors have also been used for monitoring vehicles and for assaying large crates of transuranic waste. Our Monte Carlo calculations for a new application (monitoring low-density waste for concealed plutonium) illustrate the advantages of the hollow-chamber design for detecting moderated fission sources. 9 refs., 6 figs., 2 tabs.

  16. Ionizing Radiation Detector

    DOEpatents

    Wright, Gomez W.; James, Ralph B.; Burger, Arnold; Chinn, Douglas A.

    2003-11-18

    A CdZnTe (CZT) crystal provided with a native CdO dielectric coating to reduce surface leakage currents and thereby, improve the resolution of instruments incorporating detectors using CZT crystals is disclosed. A two step process is provided for forming the dielectric coating which includes etching the surface of a CZT crystal with a solution of the conventional bromine/methanol etch treatment, and passivating the CZT crystal surface with a solution of 10 w/o NH.sub.4 F and 10 w/o H.sub.2 O.sub.2 in water after attaching electrical contacts to the crystal surface.

  17. 24 CFR 965.805 - Smoke detectors.

    Code of Federal Regulations, 2010 CFR

    2010-04-01

    ... battery-operated or hard-wired smoke detector, or such greater number as may be required by state or local... battery-operated or hard-wired smoke detectors to serve as adequate warning of fire. Public areas include... hearing-impaired residents, hard-wired smoke detectors must be connected to an alarm system designed...

  18. 24 CFR 891.555 - Smoke detectors.

    Code of Federal Regulations, 2013 CFR

    2013-04-01

    ... 24 Housing and Urban Development 4 2013-04-01 2013-04-01 false Smoke detectors. 891.555 Section... Assistance § 891.555 Smoke detectors. (a) Performance requirement. After October 30, 1992, each dwelling unit must include at least one battery-operated or hard-wired smoke detector, in proper working...

  19. 24 CFR 232.591 - Smoke detectors.

    Code of Federal Regulations, 2014 CFR

    2014-04-01

    ... 24 Housing and Urban Development 2 2014-04-01 2014-04-01 false Smoke detectors. 232.591 Section... Equipment Property Requirements § 232.591 Smoke detectors. After October 30, 1992, each occupied room must include at least one battery-operated or hard-wired smoke detector in proper working condition. If...

  20. 24 CFR 891.555 - Smoke detectors.

    Code of Federal Regulations, 2014 CFR

    2014-04-01

    ... 24 Housing and Urban Development 4 2014-04-01 2014-04-01 false Smoke detectors. 891.555 Section... Assistance § 891.555 Smoke detectors. (a) Performance requirement. After October 30, 1992, each dwelling unit must include at least one battery-operated or hard-wired smoke detector, in proper working...

  1. 24 CFR 965.805 - Smoke detectors.

    Code of Federal Regulations, 2014 CFR

    2014-04-01

    ... 24 Housing and Urban Development 4 2014-04-01 2014-04-01 false Smoke detectors. 965.805 Section... PHA-OWNED OR LEASED PROJECTS-GENERAL PROVISIONS Fire Safety § 965.805 Smoke detectors. (a) Performance... battery-operated or hard-wired smoke detector, or such greater number as may be required by state or...

  2. 24 CFR 232.591 - Smoke detectors.

    Code of Federal Regulations, 2013 CFR

    2013-04-01

    ... 24 Housing and Urban Development 2 2013-04-01 2013-04-01 false Smoke detectors. 232.591 Section... Equipment Property Requirements § 232.591 Smoke detectors. After October 30, 1992, each occupied room must include at least one battery-operated or hard-wired smoke detector in proper working condition. If...

  3. 24 CFR 242.74 - Smoke detectors.

    Code of Federal Regulations, 2012 CFR

    2012-04-01

    ... 24 Housing and Urban Development 2 2012-04-01 2012-04-01 false Smoke detectors. 242.74 Section 242... INSURANCE FOR HOSPITALS Miscellaneous Requirements § 242.74 Smoke detectors. Each occupied room must include such smoke detectors as are required by law....

  4. 24 CFR 242.74 - Smoke detectors.

    Code of Federal Regulations, 2010 CFR

    2010-04-01

    ... 24 Housing and Urban Development 2 2010-04-01 2010-04-01 false Smoke detectors. 242.74 Section 242... INSURANCE FOR HOSPITALS Miscellaneous Requirements § 242.74 Smoke detectors. Each occupied room must include such smoke detectors as are required by law....

  5. 24 CFR 242.74 - Smoke detectors.

    Code of Federal Regulations, 2011 CFR

    2011-04-01

    ... 24 Housing and Urban Development 2 2011-04-01 2011-04-01 false Smoke detectors. 242.74 Section 242... INSURANCE FOR HOSPITALS Miscellaneous Requirements § 242.74 Smoke detectors. Each occupied room must include such smoke detectors as are required by law....

  6. 24 CFR 242.74 - Smoke detectors.

    Code of Federal Regulations, 2014 CFR

    2014-04-01

    ... 24 Housing and Urban Development 2 2014-04-01 2014-04-01 false Smoke detectors. 242.74 Section 242... INSURANCE FOR HOSPITALS Miscellaneous Requirements § 242.74 Smoke detectors. Each occupied room must include such smoke detectors as are required by law....

  7. 24 CFR 242.74 - Smoke detectors.

    Code of Federal Regulations, 2013 CFR

    2013-04-01

    ... 24 Housing and Urban Development 2 2013-04-01 2013-04-01 false Smoke detectors. 242.74 Section 242... INSURANCE FOR HOSPITALS Miscellaneous Requirements § 242.74 Smoke detectors. Each occupied room must include such smoke detectors as are required by law....

  8. Development of a pixel ionization chamber for beam monitor in proton therapy

    NASA Astrophysics Data System (ADS)

    La Rosa, A.; Garella, M. A.; Attili, A.; Bourhaleb, F.; Cirio, R.; Donetti, M.; Giordanengo, S.; Givehchi, N.; Marchetto, F.; Mazza, G.; Meyroneinc, S.; Pecka, A.; Peroni, C.; Pittà, G.

    2007-03-01

    We have developed a detector to be used as monitor for proton therapy beam lines. The detector is a 2-D parallel plate ionization chamber, with the anode segmented in 1024 square pixels arranged in a 32×32 matrix. The detector characterization is presented.

  9. Wire chamber radiation detector with discharge control

    DOEpatents

    Perez-Mendez, Victor; Mulera, Terrence A.

    1984-01-01

    A wire chamber radiation detector (11) has spaced apart parallel electrodes (16) and grids (17, 18, 19) defining an ignition region (21) in which charged particles (12) or other ionizing radiations initiate brief localized avalanche discharges (93) and defining an adjacent memory region (22) in which sustained glow discharges (94) are initiated by the primary discharges (93). Conductors (29, 32) of the grids (18, 19) at each side of the memory section (22) extend in orthogonal directions enabling readout of the X-Y coordinates of locations at which charged particles (12) were detected by sequentially transmitting pulses to the conductors (29) of one grid (18) while detecting transmissions of the pulses to the orthogonal conductors (36) of the other grid (19) through glow discharges (94). One of the grids (19) bounding the memory region (22) is defined by an array of conductive elements (32) each of which is connected to the associated readout conductor (36) through a separate resistance (37). The wire chamber (11) avoids ambiguities and imprecisions in the readout of coordinates when large numbers of simultaneous or near simultaneous charged particles (12) have been detected. Down time between detection periods and the generation of radio frequency noise are also reduced.

  10. 24 CFR 200.76 - Smoke detectors.

    Code of Federal Regulations, 2014 CFR

    2014-04-01

    ... 24 Housing and Urban Development 2 2014-04-01 2014-04-01 false Smoke detectors. 200.76 Section 200.76 Housing and Urban Development Regulations Relating to Housing and Urban Development (Continued... Eligibility Requirements for Existing Projects Property Requirements § 200.76 Smoke detectors. Smoke...

  11. 24 CFR 200.76 - Smoke detectors.

    Code of Federal Regulations, 2013 CFR

    2013-04-01

    ... 24 Housing and Urban Development 2 2013-04-01 2013-04-01 false Smoke detectors. 200.76 Section 200.76 Housing and Urban Development Regulations Relating to Housing and Urban Development (Continued... Eligibility Requirements for Existing Projects Property Requirements § 200.76 Smoke detectors. Smoke...

  12. 24 CFR 200.76 - Smoke detectors.

    Code of Federal Regulations, 2012 CFR

    2012-04-01

    ... 24 Housing and Urban Development 2 2012-04-01 2012-04-01 false Smoke detectors. 200.76 Section 200.76 Housing and Urban Development Regulations Relating to Housing and Urban Development (Continued... Eligibility Requirements for Existing Projects Property Requirements § 200.76 Smoke detectors. Smoke...

  13. 24 CFR 232.591 - Smoke detectors.

    Code of Federal Regulations, 2010 CFR

    2010-04-01

    ... include at least one battery-operated or hard-wired smoke detector in proper working condition. If the... hearing-impaired persons, unless the smoke alarm is connected to a central alarm system that is...

  14. Compact ion chamber based neutron detector

    DOEpatents

    Derzon, Mark S.; Galambos, Paul C.; Renzi, Ronald F.

    2015-10-27

    A directional neutron detector has an ion chamber formed in a dielectric material; a signal electrode and a ground electrode formed in the ion chamber; a neutron absorbing material filling the ion chamber; readout circuitry which is electrically coupled to the signal and ground electrodes; and a signal processor electrically coupled to the readout circuitry. The ion chamber has a pair of substantially planar electrode surfaces. The chamber pressure of the neutron absorbing material is selected such that the reaction particle ion trail length for neutrons absorbed by the neutron absorbing material is equal to or less than the distance between the electrode surfaces. The signal processor is adapted to determine a path angle for each absorbed neutron based on the rise time of the corresponding pulse in a time-varying detector signal.

  15. Current-voltage characteristic of parallel-plane ionization chamber with inhomogeneous ionization

    NASA Astrophysics Data System (ADS)

    Stoyanov, D. G.

    2007-08-01

    The balances of particles and charges in the volume of parallel-plane ionization chamber are considered. Differential equations describing the distribution of current densities in the chamber volume are obtained. As a result of the differential equations solution an analytical form of the current-voltage characteristic of parallel-plane ionization chamber with inhomogeneous ionization in the volume is obtained.

  16. Use of a liquid ionization chamber for stereotactic radiotherapy dosimetry

    NASA Astrophysics Data System (ADS)

    Wagner, A.; Crop, F.; Lacornerie, T.; Vandevelde, F.; Reynaert, N.

    2013-04-01

    Liquid ionization chambers (LICs) offer an interesting tool in the field of small beam dosimetry, allowing better spatial resolution and reduced perturbation effects. However, some aspects remain to be addressed, such as the higher recombination and the effects from the materials of the detector. Our aim was to investigate these issues and their impact. The first step was the evaluation of the recombination effects. Measurements were performed at different SSDs to vary the dose per pulse, and the collection efficiency was obtained. The BEAMnrc code was then used to model the Cyberknife head. Finally, the liquid ionization chamber itself was modelled using the EGSnrc-based code Cavity allowing the evaluation of the influence of the volume and the chamber materials. The liquid ionization charge collection efficiency is approximately 0.98 at 1.5 mGy pulse-1, the highest dose per pulse that we have measured. Its impact on the accuracy of output factors is less than half a per cent. The detector modelling showed a significant contribution from the graphite electrode, up to 6% for the 5 mm collimator. The dependence of the average electronic mass collision stopping power of iso-octane with beam collimation is negligible and thus has no influence on output factor measurements. Finally, the volume effect reaches 5% for the small 5 mm collimator and becomes much smaller (<0.5%) for diameters above 10 mm. LICs can effectively be used for small beam relative dosimetry as long as adequate correction factors are applied, especially for the electrode and volume effects.

  17. Performance of electret ionization chambers in magnetic field.

    PubMed

    Kotrappa, P; Stieff, L R; Mengers, T F; Shull, R D

    2006-04-01

    Electret ionization chambers are widely used for measuring radon and radiation. The radiation measured includes alpha, beta, and gamma radiation. These detectors do not have any electronics and as such can be introduced into magnetic field regions. It is of interest to study the effect of magnetic fields on the performance of these detectors. Relative responses are measured with and without magnetic fields present. Quantitative responses are measured as the magnetic field is varied from 8 kA/m to 716 kA/m (100 to 9,000 gauss). No significant effect is observed for measuring alpha radiation and gamma radiation. However, a significant systematic effect is observed while measuring beta radiation from a 90Sr-Y source. Depending upon the field orientation, the relative response increased from 1.0 to 2.7 (vertical position) and decreased from 1.0 to 0.60 (horizontal position). This is explained as due to the setting up of a circular motion for the electrons by the magnetic field, which may increase or decrease the path length in air depending upon the experimental configuration. It is concluded that these ionization chambers can be used for measuring alpha (and hence radon) and gamma radiation in the range of magnetic fields studied. However, caution must be exercised if measuring beta radiation.

  18. 24 CFR 965.805 - Smoke detectors.

    Code of Federal Regulations, 2011 CFR

    2011-04-01

    ... battery-operated or hard-wired smoke detector, or such greater number as may be required by state or local... battery-operated or hard-wired smoke detectors to serve as adequate warning of fire. Public areas include... equipped with both audible and visual types of alarm devices. (2) If needed, battery-operated...

  19. 24 CFR 965.805 - Smoke detectors.

    Code of Federal Regulations, 2012 CFR

    2012-04-01

    ... battery-operated or hard-wired smoke detector, or such greater number as may be required by state or local... battery-operated or hard-wired smoke detectors to serve as adequate warning of fire. Public areas include... equipped with both audible and visual types of alarm devices. (2) If needed, battery-operated...

  20. 24 CFR 965.805 - Smoke detectors.

    Code of Federal Regulations, 2013 CFR

    2013-04-01

    ... battery-operated or hard-wired smoke detector, or such greater number as may be required by state or local... battery-operated or hard-wired smoke detectors to serve as adequate warning of fire. Public areas include... equipped with both audible and visual types of alarm devices. (2) If needed, battery-operated...

  1. NIST Ionization Chamber "A" Sample-Height Corrections.

    PubMed

    Fitzgerald, Ryan

    2012-01-01

    For over 30 years scientists in the NIST radioactivity group have been using their pressurized ionization chamber "A" (PIC "A") to make measurements of radioactivity and radioactive half-lives. We now have evidence that some of those reported measurements were incorrect due to slippage of the source positioning ring over time. The temporal change in the holder caused an error in the source-height within the chamber, which was thought to be invariant. This unaccounted-for height change caused a change in the detector response and thus a relative error in measured activity on the order of 10(-5) to 10(-3) per year, depending on the radionuclide. The drifting detector response affected calibration factors and half-life determinations. After discovering the problem, we carried out historic research and new sensitivity tests. As a result, we have created a quantitative model of the effect and have used that model to estimate corrections to some of the past measurement results from PIC "A". In this paper we report the details and results of that model. Meanwhile, we have fixed the positioning ring and are recalibrating the detector using primary measurement methods and enhanced quality control measures.

  2. Design of ionization chambers for use in teaching x-ray dosimetry

    NASA Astrophysics Data System (ADS)

    Ross, Joseph

    Ionization chambers are one of the most commonly used radiation detectors in radiation dosimetry. In this project, nine ionization chambers were constructed for use in teaching radiation dosimetry to students of health physics, medical physics, nuclear engineering, and related disciplines. The components of these detectors such as detector wall composition, type of electrode, type of leakage current guard ring, fill gas pressure, and interior conducting material differ in a systematic way to show that various parameters of ionization chamber design can affect the response of the detectors. Each of these variables was investigated using an 80 keV x-ray machine to determine detector response in terms of absorbed dose, HVL, polarity, and operating voltage. Of the components studied, wall thickness and composition was found to be the most sensitive variable. The pressure inside the chamber did have a significant effect on the amount of charge collected and the absorbed dose. The leakage current guard ring was not a critical component for this ionization chamber design.

  3. Gas amplified ionization detector for gas chromatography

    DOEpatents

    Huston, Gregg C.

    1992-01-01

    A gas-amplified ionization detector for gas chromatrography which possesses increased sensitivity and a very fast response time. Solutes eluding from a gas chromatographic column are ionized by UV photoionization of matter eluting therefrom. The detector is capable of generating easily measured voltage signals by gas amplification/multiplication of electron products resulting from the UV photoionization of at least a portion of each solute passing through the detector.

  4. Ionization detection system for aerosols

    DOEpatents

    Jacobs, Martin E.

    1977-01-01

    This invention relates to an improved smoke-detection system of the ionization-chamber type. In the preferred embodiment, the system utilizes a conventional detector head comprising a measuring ionization chamber, a reference ionization chamber, and a normally non-conductive gas triode for discharging when a threshold concentration of airborne particulates is present in the measuring chamber. The improved system utilizes a measuring ionization chamber which is modified to minimize false alarms and reductions in sensitivity resulting from changes in ambient temperature. In the preferred form of the modification, an annular radiation shield is mounted about the usual radiation source provided to effect ionization in the measuring chamber. The shield is supported by a bimetallic strip which flexes in response to changes in ambient temperature, moving the shield relative to the source so as to vary the radiative area of the source in a manner offsetting temperature-induced variations in the sensitivity of the chamber.

  5. Applying the helium ionization detector in chromatography

    NASA Technical Reports Server (NTRS)

    Gibson, E. K.; Andrawes, F. F.; Brazell, R. S.

    1981-01-01

    High noise levels and oversensitivity of helium detector make flame-ionization and thermal-conductivity detectors more suitable for chromotography. Deficiencies are eliminated by modifying helium device to operate in saturation rather than multiplication mode. Result is low background current, low noise, high stability, and high sensitivity. Detector analyzes halocarbons, hydrocarbons, hydrogen cyanide, ammonia, and inorganics without requiring expensive research-grade helium.

  6. A well-type ionization chamber geometric correction factor

    NASA Astrophysics Data System (ADS)

    Meiler, R. J.; Sibata, C. H.; Ho, A. K.; de Souza, C.; Shin, K. H.

    1996-07-01

    To correct for the influence of source configuration on the measured activity of spherical and cylindrical brachytherapy sources, a geometric correction factor was calculated for the Standard Imaging HDR-1000 well-type ionization chamber. A Fortran program modelled each source as a lattice of point sources. Because of the cylindrical symmetry of the well chamber, it could be uniquely modelled by point detectors along the perimeter of the radial plane of the detection volume. Path lengths were calculated and attenuation factors were applied to each source - detector point combination individually. The total dose rate at each detection point was found through a Sievert summation of the point source contributions. For sources with identical activities, a correction factor of was calculated, equal to the ratio of the dose rate of the cylindrical source to that of the sphere. Experimental verification using a Nuclear Associates 67-809 series cylindrical source and an Amersham spherical source yielded a correction factor of .

  7. Closed-loop pulsed helium ionization detector

    DOEpatents

    Ramsey, Roswitha S.; Todd, Richard A.

    1987-01-01

    A helium ionization detector for gas chromatography is operated in a constant current, pulse-modulated mode by configuring the detector, electrometer and a high voltage pulser in a closed-loop control system. The detector current is maintained at a fixed level by varying the frequency of fixed-width, high-voltage bias pulses applied to the detector. An output signal proportional to the pulse frequency is produced which is indicative of the charge collected for a detected species.

  8. Dosimetric characterization of a large area pixel-segmented ionization chamber.

    PubMed

    Amerio, S; Boriano, A; Bourhaleb, F; Cirio, R; Donetti, M; Fidanzio, A; Garelli, E; Giordanengo, S; Madon, E; Marchetto, F; Nastasi, U; Peroni, C; Piermattei, A; Sanz Freire, C J; Sardo, A; Trevisiol, E

    2004-02-01

    A pixel-segmented ionization chamber has been designed and built by Torino University and INFN. The detector features a 24 x 24 cm2 active area divided in 1024 independent cylindrical ionization chambers and can be read out in 500 micros without introducing dead time; the digital charge quantum can be adjusted between 100 fC and 800 fC. The sensitive volume of each single ionization chamber is 0.07 cm3. The purpose of the detector is to ease the two-dimensional (2D) verifications of fields with complex shapes and large gradients. The detector was characterized in a PMMA phantom using 60Co and 6 MV x-ray photon beams. It has shown good signal linearity with respect to dose and dose rate to water. The average sensitivity of a single ionization chamber was 2.1 nC/Gy, constant within 0.5% over one month of daily measurements. Charge collection efficiency was 0.985 at the operating polarization voltage of 400 V and 3.5 Gy/min dose rate. Tissue maximum ratio and output factor have been compared with a Farmer ionization chamber and were found in good agreement. The dose profiles have been compared with the ones obtained with an ionization chamber in water phantom for the field sizes supplied by a 3D-Line dynamic multileaf collimator. These results show that this detector can be used for 2D dosimetry of x-ray photon beams, supplying a good spatial resolution and sensibly reducing the time spent in dosimetric verification of complex radiation fields.

  9. 24 CFR 891.555 - Smoke detectors.

    Code of Federal Regulations, 2011 CFR

    2011-04-01

    ... must include at least one battery-operated or hard-wired smoke detector, in proper working condition... which case each bedroom occupied by a hearing-impaired person must have an alarm system connected to...

  10. 24 CFR 891.555 - Smoke detectors.

    Code of Federal Regulations, 2010 CFR

    2010-04-01

    ... must include at least one battery-operated or hard-wired smoke detector, in proper working condition... which case each bedroom occupied by a hearing-impaired person must have an alarm system connected to...

  11. High-resolution ionization detector and array of such detectors

    DOEpatents

    McGregor, Douglas S.; Rojeski, Ronald A.

    2001-01-16

    A high-resolution ionization detector and an array of such detectors are described which utilize a reference pattern of conductive or semiconductive material to form interaction, pervious and measurement regions in an ionization substrate of, for example, CdZnTe material. The ionization detector is a room temperature semiconductor radiation detector. Various geometries of such a detector and an array of such detectors produce room temperature operated gamma ray spectrometers with relatively high resolution. For example, a 1 cm.sup.3 detector is capable of measuring .sup.137 Cs 662 keV gamma rays with room temperature energy resolution approaching 2% at FWHM. Two major types of such detectors include a parallel strip semiconductor Frisch grid detector and the geometrically weighted trapezoid prism semiconductor Frisch grid detector. The geometrically weighted detector records room temperature (24.degree. C.) energy resolutions of 2.68% FWHM for .sup.137 Cs 662 keV gamma rays and 2.45% FWHM for .sup.60 Co 1.332 MeV gamma rays. The detectors perform well without any electronic pulse rejection, correction or compensation techniques. The devices operate at room temperature with simple commercially available NIM bin electronics and do not require special preamplifiers or cooling stages for good spectroscopic results.

  12. a Solution for Dosimetry and Quality Assurance in Imrt and Hadrontherapy:. the Pixel Ionization Chamber

    NASA Astrophysics Data System (ADS)

    Amerio, S.; Coda, S.; Nastasi, U.; Belletti, S.; Ghedi, B.; Boriano, A.; Cirio, R.; Luparia, A.; Marchetto, F.; Peroni, C.; Sanz Freire, C. J.; Donetti, M.; Madon, E.; Trevisiol, E.; Urgesi, A.

    2002-11-01

    The new radiotherapy techniques require new detectors to monitor and measure the clinical field. The Intensity Modulated Radiation Therapy (IMRT) techniques like step and shoot, sliding window, dynamic wedge or scanning beam add the time variable to the treatment field. In this case the water phantom with a single ionization chamber moving inside the field needs very long measurement time. Linear arrays of ionization chambers or diodes measure the field only along a line. 2D detectors like radiographic or gafchromic film are not suitable to be used as on line detectors. We have developed, built and tested an ionization chamber segmented in pixels that measure the dose in a plane at several points. Every channel has a dedicated electronic chain that digitizes the collected charge and data from all the channels are sent to the computer that performs the data acquisition. One read out cycle is very fast allowing to measure in real time the fluency and the shape of the field. The chamber can be used in two different ways, as monitor chamber and as relative dosemeter. A description of the detector, the electronics, and test results with both photon and hadron beams will be reported.

  13. High pressure xenon ionization detector

    DOEpatents

    Markey, J.K.

    1989-11-14

    A method is provided for detecting ionization comprising allowing particles that cause ionization to contact high pressure xenon maintained at or near its critical point and measuring the amount of ionization. An apparatus is provided for detecting ionization, the apparatus comprising a vessel containing a ionizable medium, the vessel having an inlet to allow high pressure ionizable medium to enter the vessel, a means to permit particles that cause ionization of the medium to enter the vessel, an anode, a cathode, a grid and a plurality of annular field shaping rings, the field shaping rings being electrically isolated from one another, the anode, cathode, grid and field shaping rings being electrically isolated from one another in order to form an electric field between the cathode and the anode, the electric field originating at the anode and terminating at the cathode, the grid being disposed between the cathode and the anode, the field shaping rings being disposed between the cathode and the grid, the improvement comprising the medium being xenon and the vessel being maintained at a pressure of 50 to 70 atmospheres and a temperature of 0 to 30 C. 2 figs.

  14. High pressure xenon ionization detector

    DOEpatents

    Markey, John K.

    1989-01-01

    A method is provided for detecting ionization comprising allowing particles that cause ionization to contact high pressure xenon maintained at or near its critical point and measuring the amount of ionization. An apparatus is provided for detecting ionization, the apparatus comprising a vessel containing a ionizable medium, the vessel having an inlet to allow high pressure ionizable medium to enter the vessel, a means to permit particles that cause ionization of the medium to enter the vessel, an anode, a cathode, a grid and a plurality of annular field shaping rings, the field shaping rings being electrically isolated from one another, the anode, cathode, grid and field shaping rings being electrically isolated from one another in order to form an electric field between the cathode and the anode, the electric field originating at the anode and terminating at the cathode, the grid being disposed between the cathode and the anode, the field shaping rings being disposed between the cathode and the grid, the improvement comprising the medium being xenon and the vessel being maintained at a pressure of 50 to 70 atmospheres and a temperature of 0.degree. to 30.degree. C.

  15. Construction of a fast ionization chamber for high-rate particle identification

    NASA Astrophysics Data System (ADS)

    Chae, K. Y.; Ahn, S.; Bardayan, D. W.; Chipps, K. A.; Manning, B.; Pain, S. D.; Peters, W. A.; Schmitt, K. T.; Smith, M. S.; Strauss, S. Y.

    2014-07-01

    A new gas-filled ionization chamber for high count rate particle identification has been constructed and commissioned at the Holifield Radioactive Ion Beam Facility (HRIBF) at Oak Ridge National Laboratory (ORNL). To enhance the response time of the ionization chamber, a design utilizing a tilted entrance window and tilted electrodes was adopted, which is modified from an original design by Kimura et al. [1]. A maximum counting rate of 700 , 000 particles per second has been achieved. The detector has been used for several radioactive beam measurements performed at the HRIBF.

  16. Super-resolution non-parametric deconvolution in modelling the radial response function of a parallel plate ionization chamber.

    PubMed

    Kulmala, A; Tenhunen, M

    2012-11-07

    The signal of the dosimetric detector is generally dependent on the shape and size of the sensitive volume of the detector. In order to optimize the performance of the detector and reliability of the output signal the effect of the detector size should be corrected or, at least, taken into account. The response of the detector can be modelled using the convolution theorem that connects the system input (actual dose), output (measured result) and the effect of the detector (response function) by a linear convolution operator. We have developed the super-resolution and non-parametric deconvolution method for determination of the cylinder symmetric ionization chamber radial response function. We have demonstrated that the presented deconvolution method is able to determine the radial response for the Roos parallel plate ionization chamber with a better than 0.5 mm correspondence with the physical measures of the chamber. In addition, the performance of the method was proved by the excellent agreement between the output factors of the stereotactic conical collimators (4-20 mm diameter) measured by the Roos chamber, where the detector size is larger than the measured field, and the reference detector (diode). The presented deconvolution method has a potential in providing reference data for more accurate physical models of the ionization chamber as well as for improving and enhancing the performance of the detectors in specific dosimetric problems.

  17. RADIATION MONITOR CONTAINING TWO CONCENTRIC IONIZATION CHAMBERS AND MEANS FOR INSULATING THE SEPARATE CHAMBERS

    DOEpatents

    Braestrup, C.B.; Mooney, R.T.

    1964-01-21

    This invention relates to a portable radiation monitor containing two concentric ionization chambers which permit the use of standard charging and reading devices. It is particularly adapted as a personnel x-ray dosimeter and to this end comprises a small thin walled, cylindrical conductor forming an inner energy dependent chamber, a small thin walled, cylindrical conductor forming an outer energy independent chamber, and polymeric insulation means which insulates said chambers from each other and holds the chambers together with exposed connections in a simple, trouble-free, and compact assembly substantially without variation in directional response. (AEC)

  18. Quantification of static magnetic field effects on radiotherapy ionization chambers

    NASA Astrophysics Data System (ADS)

    Agnew, J.; O’Grady, F.; Young, R.; Duane, S.; Budgell, G. J.

    2017-03-01

    Integrated magnetic resonance (MR) imaging and radiotherapy (RT) delivery machines are currently being developed, with some already in clinical use. It is anticipated that the strong magnetic field used in some MR-RT designs will have a significant impact on routine measurements of dose in the MR-linac performed using ionization chambers, which provide traceability back to a primary standard definition of dose. In particular, the presence of small air gaps around ionization chambers may introduce unacceptably high uncertainty into these measurements. In this study, we investigate and quantify the variation attributable to air gaps for several routinely-used cylindrical ionization chambers in a magnetic field, as well as the effect of the magnetic field alone on the response of the chambers. The measurements were performed in a Co-60 beam, while the ionization chambers were positioned in custom-made Perspex phantoms between the poles of an electromagnet, which was capable of generating magnetic fields of up to 2 T field strength, although measurements were focused around 1.5 T. When an asymmetric air gap was rotated at cardinal angles around the ionization chambers investigated here, variation of up to 8.5  ±  0.2 percentage points (PTW 31006 chamber) was observed in an applied magnetic field of 1.5 T. The minimum peak-to-peak variation was 1.1  ±  0.1% (Exradin A1SL). When the same experiment was performed with a well-defined air gap of known position using the PTW 30013 chamber, a variation of 3.8  ±  0.2% was observed. When water was added to the phantom cavity to eliminate all air gaps, the variation for the PTW 30013 was reduced to 0.2  ±  0.01%.

  19. Quantification of static magnetic field effects on radiotherapy ionization chambers.

    PubMed

    Agnew, J; O'Grady, F; Young, R; Duane, S; Budgell, G J

    2017-03-07

    Integrated magnetic resonance (MR) imaging and radiotherapy (RT) delivery machines are currently being developed, with some already in clinical use. It is anticipated that the strong magnetic field used in some MR-RT designs will have a significant impact on routine measurements of dose in the MR-linac performed using ionization chambers, which provide traceability back to a primary standard definition of dose. In particular, the presence of small air gaps around ionization chambers may introduce unacceptably high uncertainty into these measurements. In this study, we investigate and quantify the variation attributable to air gaps for several routinely-used cylindrical ionization chambers in a magnetic field, as well as the effect of the magnetic field alone on the response of the chambers. The measurements were performed in a Co-60 beam, while the ionization chambers were positioned in custom-made Perspex phantoms between the poles of an electromagnet, which was capable of generating magnetic fields of up to 2 T field strength, although measurements were focused around 1.5 T. When an asymmetric air gap was rotated at cardinal angles around the ionization chambers investigated here, variation of up to 8.5  ±  0.2 percentage points (PTW 31006 chamber) was observed in an applied magnetic field of 1.5 T. The minimum peak-to-peak variation was 1.1  ±  0.1% (Exradin A1SL). When the same experiment was performed with a well-defined air gap of known position using the PTW 30013 chamber, a variation of 3.8  ±  0.2% was observed. When water was added to the phantom cavity to eliminate all air gaps, the variation for the PTW 30013 was reduced to 0.2  ±  0.01%.

  20. Fiber optic linear smoke fire detector

    NASA Astrophysics Data System (ADS)

    Kulakov, Sergei V.; Moskaletz, Oleg D.; Preslenev, Leonid N.; Shabardin, Alexander N.

    2001-11-01

    A global and versatile problem of fire and environmental safety is formulated. It is pointed out that one of the main ways to solve this problem is the development of equipment for early fire detection. The results of the development and study of a smoke fiber optic fire detector are presented. Such detector is absolutely explosion-safe and immune to increased radiation level and aggressive chemical environment.

  1. A position-sensitive twin ionization chamber for fission fragment and prompt neutron correlation experiments

    NASA Astrophysics Data System (ADS)

    Göök, A.; Geerts, W.; Hambsch, F.-J.; Oberstedt, S.; Vidali, M.; Zeynalov, Sh.

    2016-09-01

    A twin position-sensitive Frisch grid ionization chamber, intended as a fission fragment detector in experiments to study prompt fission neutron correlations with fission fragment properties, is presented. Fission fragment mass and energies are determined by means of the double kinetic energy technique, based on conservation of mass and linear momentum. The position sensitivity is achieved by replacing each anode plate in the standard twin ionization chamber by a wire plane and a strip anode, both readout by means of resistive charge division. This provides information about the fission axis orientation, which is necessary to reconstruct the neutron emission process in the fully accelerated fragment rest-frame. The energy resolution compared to the standard twin ionization chamber is found not to be affected by the modification. The angular resolution of the detector relative to an arbitrarily oriented axis is better than 7° FWHM. Results on prompt fission neutron angular distributions in 235U(n,f) obtained with the detector in combination with an array of neutron scintillation detectors is presented as a proof of principle.

  2. Absolute dosimetry on a dynamically scanned sample for synchrotron radiotherapy using graphite calorimetry and ionization chambers

    NASA Astrophysics Data System (ADS)

    Lye, J. E.; Harty, P. D.; Butler, D. J.; Crosbie, J. C.; Livingstone, J.; Poole, C. M.; Ramanathan, G.; Wright, T.; Stevenson, A. W.

    2016-06-01

    The absolute dose delivered to a dynamically scanned sample in the Imaging and Medical Beamline (IMBL) on the Australian Synchrotron was measured with a graphite calorimeter anticipated to be established as a primary standard for synchrotron dosimetry. The calorimetry was compared to measurements using a free-air chamber (FAC), a PTW 31 014 Pinpoint ionization chamber, and a PTW 34 001 Roos ionization chamber. The IMBL beam height is limited to approximately 2 mm. To produce clinically useful beams of a few centimetres the beam must be scanned in the vertical direction. In practice it is the patient/detector that is scanned and the scanning velocity defines the dose that is delivered. The calorimeter, FAC, and Roos chamber measure the dose area product which is then converted to central axis dose with the scanned beam area derived from Monte Carlo (MC) simulations and film measurements. The Pinpoint chamber measures the central axis dose directly and does not require beam area measurements. The calorimeter and FAC measure dose from first principles. The calorimetry requires conversion of the measured absorbed dose to graphite to absorbed dose to water using MC calculations with the EGSnrc code. Air kerma measurements from the free air chamber were converted to absorbed dose to water using the AAPM TG-61 protocol. The two ionization chambers are secondary standards requiring calibration with kilovoltage x-ray tubes. The Roos and Pinpoint chambers were calibrated against the Australian primary standard for air kerma at the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA). Agreement of order 2% or better was obtained between the calorimetry and ionization chambers. The FAC measured a dose 3-5% higher than the calorimetry, within the stated uncertainties.

  3. Absolute dosimetry on a dynamically scanned sample for synchrotron radiotherapy using graphite calorimetry and ionization chambers.

    PubMed

    Lye, J E; Harty, P D; Butler, D J; Crosbie, J C; Livingstone, J; Poole, C M; Ramanathan, G; Wright, T; Stevenson, A W

    2016-06-07

    The absolute dose delivered to a dynamically scanned sample in the Imaging and Medical Beamline (IMBL) on the Australian Synchrotron was measured with a graphite calorimeter anticipated to be established as a primary standard for synchrotron dosimetry. The calorimetry was compared to measurements using a free-air chamber (FAC), a PTW 31 014 Pinpoint ionization chamber, and a PTW 34 001 Roos ionization chamber. The IMBL beam height is limited to approximately 2 mm. To produce clinically useful beams of a few centimetres the beam must be scanned in the vertical direction. In practice it is the patient/detector that is scanned and the scanning velocity defines the dose that is delivered. The calorimeter, FAC, and Roos chamber measure the dose area product which is then converted to central axis dose with the scanned beam area derived from Monte Carlo (MC) simulations and film measurements. The Pinpoint chamber measures the central axis dose directly and does not require beam area measurements. The calorimeter and FAC measure dose from first principles. The calorimetry requires conversion of the measured absorbed dose to graphite to absorbed dose to water using MC calculations with the EGSnrc code. Air kerma measurements from the free air chamber were converted to absorbed dose to water using the AAPM TG-61 protocol. The two ionization chambers are secondary standards requiring calibration with kilovoltage x-ray tubes. The Roos and Pinpoint chambers were calibrated against the Australian primary standard for air kerma at the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA). Agreement of order 2% or better was obtained between the calorimetry and ionization chambers. The FAC measured a dose 3-5% higher than the calorimetry, within the stated uncertainties.

  4. Pulse mode readout techniques for use with non-gridded industrial ionization chambers

    SciTech Connect

    Popov, Vladimir E.; Degtiarenko, Pavel V.

    2011-10-01

    Highly sensitive readout technique for precision long-term radiation measurements has been developed and tested in the Radiation Control Department at Jefferson Lab. The new electronics design is used to retrieve ionization data in a pulse mode. The dedicated data acquisition system works with M=Audio Audiophile 192 High-Definition 24-bit/192 kHz audio cards, taking data in continuous waveform recording mode. The on-line data processing algorithms extract signals of the ionization events from the data flow and measure the ionization value for each event. Two different ion chambers are evaluated. The first is a Reuter-Stokes Argon-filled (at 25 atm) High Pressure Ionization Chamber (HPIC), commonly used as a detector part in many GE Reuter-Stokes instruments of the RSS series. The second is a VacuTec Model 70181, 5 atm Xenon-filled ionization chamber. Results for both chambers indicate that the techniques allow using industrial ICs for high sensitivity and precision long-term radiation measurements, while at the same time providing information about spectral characteristics of the radiation fields.

  5. Hard disk drive based microsecond x-ray chopper for characterization of ionization chambers and photodiodes

    SciTech Connect

    Müller, O. Lützenkirchen-Hecht, D.; Frahm, R.

    2015-03-15

    A fast X-ray chopper capable of producing ms long X-ray pulses with a typical rise time of few μs was realized. It is ideally suited to investigate the temporal response of X-ray detectors with response times of the order of μs to ms, in particular, any kind of ionization chambers and large area photo diodes. The drive mechanism consists of a brushless DC motor and driver electronics from a common hard disk drive, keeping the cost at an absolute minimum. Due to its simple construction and small dimensions, this chopper operates at home lab based X-ray tubes and synchrotron radiation sources as well. The dynamics of the most important detectors used in time resolved X-ray absorption spectroscopy, namely, ionization chambers and Passivated Implanted Planar Silicon photodiodes, were investigated in detail. The results emphasize the applicability of this X-ray chopper.

  6. Hard disk drive based microsecond X-ray chopper for characterization of ionization chambers and photodiodes.

    PubMed

    Müller, O; Lützenkirchen-Hecht, D; Frahm, R

    2015-03-01

    A fast X-ray chopper capable of producing ms long X-ray pulses with a typical rise time of few μs was realized. It is ideally suited to investigate the temporal response of X-ray detectors with response times of the order of μs to ms, in particular, any kind of ionization chambers and large area photo diodes. The drive mechanism consists of a brushless DC motor and driver electronics from a common hard disk drive, keeping the cost at an absolute minimum. Due to its simple construction and small dimensions, this chopper operates at home lab based X-ray tubes and synchrotron radiation sources as well. The dynamics of the most important detectors used in time resolved X-ray absorption spectroscopy, namely, ionization chambers and Passivated Implanted Planar Silicon photodiodes, were investigated in detail. The results emphasize the applicability of this X-ray chopper.

  7. Hard disk drive based microsecond x-ray chopper for characterization of ionization chambers and photodiodes

    NASA Astrophysics Data System (ADS)

    Müller, O.; Lützenkirchen-Hecht, D.; Frahm, R.

    2015-03-01

    A fast X-ray chopper capable of producing ms long X-ray pulses with a typical rise time of few μs was realized. It is ideally suited to investigate the temporal response of X-ray detectors with response times of the order of μs to ms, in particular, any kind of ionization chambers and large area photo diodes. The drive mechanism consists of a brushless DC motor and driver electronics from a common hard disk drive, keeping the cost at an absolute minimum. Due to its simple construction and small dimensions, this chopper operates at home lab based X-ray tubes and synchrotron radiation sources as well. The dynamics of the most important detectors used in time resolved X-ray absorption spectroscopy, namely, ionization chambers and Passivated Implanted Planar Silicon photodiodes, were investigated in detail. The results emphasize the applicability of this X-ray chopper.

  8. High-resolution ion pulse ionization chamber with air filling for the 222Rn decays detection

    NASA Astrophysics Data System (ADS)

    Gavrilyuk, Yu. M.; Gangapshev, A. M.; Gezhaev, A. M.; Etezov, R. A.; Kazalov, V. V.; Kuzminov, V. V.; Panasenko, S. I.; Ratkevich, S. S.; Tekueva, D. A.; Yakimenko, S. P.

    2015-11-01

    The construction and characteristics of the cylindrical ion pulse ionization chamber (CIPIC) with a working volume of 3.2 L are described. The chamber is intended to register α-particles from the 222Rn and its daughter's decays in the filled air sample. The detector is less sensitive to electromagnetic pick-ups and mechanical noises. The digital pulse processing method is proposed to improve the energy resolution of the ion pulse ionization chamber. An energy resolution of 1.6% has been achieved for the 5.49 MeV α-line. The dependence of the energy resolution on high voltage and working media pressure has been investigated and the results are presented.

  9. Assessment of radiation doses from residential smoke detectors that contain americium-241

    SciTech Connect

    O'Donnell, F.R.; Etnier, E.L.; Holton, G.A.; Travis, C.C.

    1981-10-01

    External dose equivalents and internal dose commitments were estimated for individuals and populations from annual distribution, use, and disposal of 10 million ionization chamber smoke detectors that contain 110 kBq (3 ..mu..Ci) americium-241 each. Under exposure scenarios developed for normal distribution, use, and disposal using the best available information, annual external dose equivalents to average individuals were estimated to range from 4 fSv (0.4 prem) to 20 nSv (2 ..mu..rem) for total body and from 7 fSv to 40 nSv for bone. Internal dose commitments to individuals under post disposal scenarios were estimated to range from 0.006 to 80 ..mu..Sv (0.0006 to 8 mrem) to total body and from 0.06 to 800 ..mu..Sv to bone. The total collective dose (the sum of external dose equivalents and 50-year internal dose commitments) for all individuals involved with distribution, use, or disposal of 10 million smoke detectors was estimated to be about 0.38 person-Sv (38 person-rem) to total body and 00 ft/sup 2/).

  10. EML pulse ionization chamber systems for /sup 222/Rn measurements

    SciTech Connect

    Fisenne, I M; Keller, H W

    1985-03-01

    Radon measurements have been performed with pulse ionization chambers at the Environmental Measurements Laboratory (EML) for over 35 years. This report describes the evolution of radon measurement systems, with emphasis on the continuous quality control efforts at EML. 38 refs., 3 figs., 3 tabs.

  11. High-rate axial-field ionization chamber for particle identification of radioactive beams

    NASA Astrophysics Data System (ADS)

    Vadas, J.; Singh, Varinderjit; Visser, G.; Alexander, A.; Hudan, S.; Huston, J.; Wiggins, B. B.; Chbihi, A.; Famiano, M.; Bischak, M. M.; deSouza, R. T.

    2016-11-01

    The design, construction and performance characteristics of a simple axial-field ionization chamber suitable for identifying ions in a radioactive beam are presented. Optimized for use with low-energy radioactive beams (< 5 MeV / A) the detector presents only three 0.5 μm/cm2 foils to the beam in addition to the detector gas. A fast charge sensitive amplifier (CSA) integrated into the detector design is also described. Coupling this fast CSA to the axial field ionization chamber produces an output pulse with a risetime of 60-70 ns and a fall time of 100 ns, making the detector capable of sustaining a relatively high rate and providing a time resolution of 6-8 ns. Tests with an α source establish the detector energy resolution as ∼ 8 % for an energy deposit of ∼3.5 MeV. The energy resolution with beams of 2.5 and 4.0 MeV/A 39K ions and the dependence of the energy resolution on beam intensity is measured. At an instantaneous rate of 3×105 ions/s the energy resolution has degraded to 14% with a pileup of 12%. The good energy resolution of this detector at rates up to 3×105 ions/s makes it an effective tool in the characterization of low-energy radioactive beams.

  12. Ionization chamber gradient effects in nonstandard beam configurations

    SciTech Connect

    Bouchard, Hugo; Seuntjens, Jan; Carrier, Jean-Francois; Kawrakow, Iwan

    2009-10-15

    Purpose: For the purpose of nonstandard beam reference dosimetry, the current concept of reporting absorbed dose at a point in water located at a representative position in the chamber volume is investigated in detail. As new nonstandard beam reference dosimetry protocols are under development, an evaluation of the role played by the definition of point of measurement could lead to conceptual improvements prior to establishing measurement procedures. Methods: The present study uses the current definition of reporting absorbed dose to calculate ionization chamber perturbation factors for two cylindrical chamber models (Exradin A12 and A14) using the Monte Carlo method. The EGSnrc based user-code EGS lowbar chamber is used to calculate chamber dose responses of 14 IMRT beams chosen to cause considerable dose gradients over the chamber volume as previously used by Bouchard and Seuntjens [''Ionization chamber-based reference dosimetry of intensity modulated radiation beams,'' Med. Phys. 31(9), 2454-5465 (2004)]. Results: The study shows conclusively the relative importance of each physical effect involved in the nonstandard beam correction factors of 14 IMRT beams. Of all correction factors involved in the dosimetry of the beams studied, the gradient perturbation correction factor has the highest magnitude, on average, 11% higher compared to reference conditions for the Exradin A12 chamber and about 5% higher for the Extradin A14 chamber. Other perturbation correction factors (i.e., P{sub wall}, P{sub stem}, and P{sub cel}) are, on average, less than 0.8% different from reference conditions for the chambers and beams studied. The current approach of reporting measured absorbed dose at a point in water coinciding with the location of the centroid of the chamber is the main factor responsible for large correction factors in nonstandard beam deliveries (e.g., intensity modulated radiation therapy) reported in literature. Conclusions: To reduce or eliminate the magnitude

  13. Wire-chamber radiation detector with discharge control

    DOEpatents

    Perez-Mendez, V.; Mulera, T.A.

    1982-03-29

    A wire chamber; radiation detector has spaced apart parallel electrodes and grids defining an ignition region in which charged particles or other ionizing radiations initiate brief localized avalanche discharges and defining an adjacent memory region in which sustained glow discharges are initiated by the primary discharges. Conductors of the grids at each side of the memory section extend in orthogonal directions enabling readout of the X-Y coordinates of locations at which charged particles were detected by sequentially transmitting pulses to the conductors of one grid while detecting transmissions of the pulses to the orthogonal conductors of the other grid through glow discharges. One of the grids bounding the memory region is defined by an array of conductive elements each of which is connected to the associated readout conductor through a separate resistance. The wire chamber avoids ambiguities and imprecisions in the readout of coordinates when large numbers of simultaneous or; near simultaneous charged particles have been detected. Down time between detection periods and the generation of radio frequency noise are also reduced.

  14. High-rate axial-field ionization chamber for particle identification of Radioactive beams

    NASA Astrophysics Data System (ADS)

    Desouza, Romualdo; Vadas, Justin; Singh, Varinderjit; Visser, G.; Alexander, A.; Hudan, S.; Huston, J.; Wiggins, B.; Chbihi, A.; Famiano, M.; Bischak, M.

    2017-01-01

    The design, construction and performance characteristics of a simple axial-field ionization chamber suitable for identifying ions in a radioactive beam are presented. The detector is optimized for use with low-energy radioactive beams (<) 5 MeV/A. A fast charge sensitive amplifier (CSA) integrated into the detector design is also described. Coupling this fast CSA to the axial field ionization chamber produces an output pulse with a rise-time of 60 to 70 ns and a fall time of 100 ns, making the detector capable of sustaining a relatively high rate while providing a time resolution of 6 to 8 ns. Tests with an α source establish the detector energy resolution as 8 % for an energy deposit of 3.5 MeV. Beam tests indicate that the detector is an effective tool for the characterization of low-energy radioactive beams at beam intensities up to 3 x 105 ions/s. Supported by the U.S. DOE under Award # DE-FG02-88ER-40404 and the NSF under Grant No. 1342962.

  15. Neutron Dosimetry Using a Tissue-Equivalent Ionization Chamber.

    DTIC Science & Technology

    1980-05-01

    procedures are described, and correction factors discussed. /’ On a montg et v~rifiA un systtme de niesure d’oZz la dose ou le kerma tissulaire neutronique ...tube 3.8 cm in diameter. At the same time, it is advantageous to keep the active volume as large as possible in order to maximize sensitivity. The...CHAMBER The construction of the FWT TE Ionization chamber is Illustrated in figure 1. The active volume lies between the central electrode and the

  16. Monte Carlo Simulation in the Optimization of a Free-Air Ionization Chamber for Dosimetric Control in Medical Digital Radiography

    SciTech Connect

    Leyva, A.; Pinera, I.; Abreu, Y.; Cruz, C. M.; Montano, L. M.

    2008-08-11

    During the earliest tests of a free-air ionization chamber a poor response to the X-rays emitted by several sources was observed. Then, the Monte Carlo simulation of X-rays transport in matter was employed in order to evaluate chamber behavior as X-rays detector. The photons energy deposition dependence with depth and its integral value in all active volume were calculated. The obtained results reveal that the designed device geometry is feasible to be optimized.

  17. Strip Ionization Chamber as Beam Monitor in the Proton Therapy Eye Treatment

    NASA Astrophysics Data System (ADS)

    Marchetto, F.; Cirio, R.; Garella, M. A.; Giordanengo, S.; Boriano, A.; Givehchi, N.; La Rosa, A.; Peroni, C.; Donetti, M.; Bourhaleb, F.; Pitta', G.; Cirrone, G. A. P.; Cuttone, G.; Raffaele, L.; Sabini, M. G.; Valastro, L.

    2006-04-01

    Since spring 2002, ocular pathologies have been treated in Catania at the Centro di AdroTerapia e Applicazioni Nucleari Avanzate (CATANA) within a collaboration between INFN Laboratori Nazionali del Sud (LNS), Physics Department, Ophthalmology Institute, Radiology Institute of the Catania University and CSFNSM Catania. A beam line from a 62 MeV Superconducting Cyclotron is used to treat shallow tumors. The beam is conformed to the tumor shape with a passive delivery system. A detector system has been developed in collaboration with INFN-Torino to be used as real time beam monitor. The detector, placed upstream of the patient collimator, consists of two parallel plate ionization chambers with the anode segmented in strips. Each anode is made of 0.5 mm-wide 256 strips corresponding to (12.8 × 12.8) cm2 sensitive area. With the two strip ionization chambers one can measure the relevant beam parameters during treatment to probe both asymmetry and flatness. In the test carried out at CATANA the detector has been used under different and extreme beam conditions. Preliminary results are given for profiles and skewness, together with a comparison with reference detectors.

  18. Nuclear signal simulation applied to gas ionizing chambers

    SciTech Connect

    Coulon, Romain; Dumazert, Jonathan

    2015-07-01

    Particle transport codes used in detector simulation allow the calculation of the energy deposited by charged particles produced following an interaction. The pulses temporal shaping is more and more used in nuclear measurement into pulse shape analysis techniques. A model is proposed in this paper to simulate the pulse temporal shaping and the associated noise level thanks to the output track file PTRAC provides by Monte-Carlo particle transport codes. The model has been dedicated to ion chambers and more especially for High Pressure Xenon chambers HPXe where the pulse shape analysis can resolve some issues regarding with this technology as the ballistic deficit phenomenon. The model is fully described and an example is presented as a validation of such full detector simulation. (authors)

  19. An evaluation of ionization chambers for the relative dosimetry of kilovoltage x-ray beams

    SciTech Connect

    Hill, Robin; Mo Zhao; Haque, Mamoon; Baldock, Clive

    2009-09-15

    In this work, the authors have evaluated ten different ionization chambers for the relative dosimetry of kilovoltage x-ray beams in the energy range of 50-280 kVp. Percentage depth doses in water and relative detector response (in Solid Water and in air) were measured for each of the x-ray beams studied using a number of chambers. Measured depth dose data were compared with Monte Carlo calculated depth doses using the EGSnrc Monte Carlo package and the BEAMnrc user code. The accuracy of the phase space files generated by BEAMnrc was verified by calculating the half-value layer and comparing with the measured half-value layer of each x-ray beam. The results indicate that the Advanced Markus, Markus, NACP, and Roos parallel plate ionization chambers were suitable for the measurement of depth dose data in this beam quality range with an uncertainty of less than 3%, including in the regions close to the water surface. While the relative detector response of the Farmer and scanning thimble chambers exhibited a better energy response, they were not suitable for depth dose measurements in the first 5 mm below the water surface with differences of up to 12% in the surface dose measurement for the 50 kVp x-ray beam. These differences were due to dose artifacts generated by the chamber size and the dose gradient. However, at depths greater than 5 mm, the Farmer and thimble scanning chambers gave uncertainties of less than 3% for the depth dose measurements for all beam energies. The PTW PinPoint 31006 chamber was found to give varying dose differences of up to 8% depending on the x-ray beam energy; this was attributed to the steel central electrode. The authors recommend that one of the parallel plate ionization chambers investigated be used to determine depth dose data for kilovoltage x-ray beams in the energy range studied and give correct dose information close to the surface and at depth in the water phantom.

  20. Directional Dark Matter Detector Prototype (Time Projection Chamber)

    NASA Astrophysics Data System (ADS)

    Oliver-Mallory, Kelsey; Garcia-Sciveres, Maurice; Kadyk, John; Lopex-Thibodeaux, Mayra

    2013-04-01

    The time projection chamber is a mature technology that has emerged as a promising candidate for the directional detection of the WIMP particle. In order to utilize this technology in WIMP detection, the operational parameters must be chosen in the non-ideal regime. A prototype WIMP detector with a 10cm field cage, double GEM amplification, and ATLAS FEI3 pixel chip readout was constructed for the purpose of investigating effects of varying gas pressure in different gas mixtures. The rms radii of ionization clusters of photoelectrons caused by X-rays from a Fe-55 source were measured for several gas pressures between 760torr and 99torr in Ar(70)/ CO2(30), CF4, He(80)/Isobutane(20), and He(80)/CF4(20) mixtures. Average radii were determined from distributions of the data for each gas mixture and pressure, and revealed a negative correlation between pressure and radius in Ar(70)/CO2(30) and He(80)/Isobutane(20) mixtures. Investigation of the pressure-radius measurements are in progress using distributions of photoelectron and auger electron practical ranges (Univ. of Pisa) and diffusion, using the Garfield Monte Carlo program.

  1. Comparison of ionization chamber efficiencies for activity measurements.

    PubMed

    Schrader, H; Svec, A

    2004-01-01

    The calibration of ionization chamber measuring systems in terms of activity is described. The energy-dependent efficiency curves of three chambers at the Bureau International des Poids et Mesures, the National Physical Laboratory and the Physikalisch-Technische Bundesanstalt are determined and compared using a fitting procedure for the experimental radionuclide efficiencies by the Microsoft (MS) EXCEL Solver program. An estimation of the uncertainty of the efficiency curves and the deviations of experimental and calculated radionuclide efficiencies are given. By this fitting method, discrepancies in the efficiency determination can be detected at a level of about one percent. Systematic deviations entering into the calculations either from emission probabilities per decay or from absolute activity standardization are discussed.

  2. Amplitude distribution of ionization jerks in ionization-chamber ASK-1 according long-term measurements

    NASA Astrophysics Data System (ADS)

    Timofeev, Vladislav

    2016-07-01

    As part of the Yakut complex systems by measuring the intensity of cosmic rays has a unique device spherical - ionization chamber ASK-1 with a lead screen thickness of 12 cm. The camera allows you to explore the physical characteristics of the so-called "ionization jerks " - sharp increases ionization current caused by the passage through the device much ionizing particles of cosmic origin. Due to a large increase in current nuclear cascade "showers", formed mainly by particles of cosmic rays in the camera screen. Over the entire period of observation (50 years old) camera ASK-1 was registered 59125 aftershocks. Their nature and properties still does not sufficiently studied, especially in medium and large amplitudes.

  3. SU-E-T-525: Ionization Chamber Perturbation in Flattening Filter Free Beams

    SciTech Connect

    Czarnecki, D; Voigts-Rhetz, P von; Zink, K

    2015-06-15

    Purpose: Changing the characteristic of a photon beam by mechanically removing the flattening filter may impact the dose response of ionization chambers. Thus, perturbation factors of cylindrical ionization chambers in conventional and flattening filter free photon beams were calculated by Monte Carlo simulations. Methods: The EGSnrc/BEAMnrc code system was used for all Monte Carlo calculations. BEAMnrc models of nine different linear accelerators with and without flattening filter were used to create realistic photon sources. Monte Carlo based calculations to determine the fluence perturbations due to the presens of the chambers components, the different material of the sensitive volume (air instead of water) as well as the volume effect were performed by the user code egs-chamber. Results: Stem, central electrode, wall, density and volume perturbation factors for linear accelerators with and without flattening filter were calculated as a function of the beam quality specifier TPR{sub 20/10}. A bias between the perturbation factors as a function of TPR{sub 20/10} for flattening filter free beams and conventional linear accelerators could not be observed for the perturbations caused by the components of the ionization chamber and the sensitive volume. Conclusion: The results indicate that the well-known small bias between the beam quality correction factor as a function of TPR20/10 for the flattening filter free and conventional linear accelerators is not caused by the geometry of the detector but rather by the material of the sensitive volume. This suggest that the bias for flattening filter free photon fields is only caused by the different material of the sensitive volume (air instead of water)

  4. Bubble chamber as a trace chemical detector.

    PubMed

    Luo, X; McCreary, E I; Atencio, J H; McCown, A W; Sander, R K

    1998-08-20

    A novel concept for trace chemical analysis in liquids has been demonstrated. The technique utilizes light absorption in a superheated liquid. Although a superheated liquid is thermodynamically unstable, a high degree of superheating can be dynamically achieved for a short period of time. During this time the superheated liquid is extremely sensitive to boiling at nucleation sites produced by energy deposition. Observation of bubbles in the superheated liquid in some sense provides amplification of the initial energy deposition. Bubble chambers containing superheated liquids have been used to detect energetic particles; now a bubble chamber is used to detect a trace chemical in superheated liquid propane by observing bubble formation initiated by optical absorption. Crystal violet is used as a test case and can be detected at the subpart-per-10(12) level by using a Nd:YAG laser. The mechanism for bubble formation and ideas for further improvement are discussed.

  5. Analytical form of current-voltage characteristic of parallel-plane, cylindrical and spherical ionization chambers with homogeneous ionization

    NASA Astrophysics Data System (ADS)

    Stoyanov, D. G.

    2007-11-01

    The elementary processes taking place in the formation of charged particles and their flow in parallel-plane, cylindrical and spherical geometry cases of ionization chamber are considered. On the basis of particles and charges balance a differential equation describing the distribution of current densities in the ionization chamber volume is obtained. As a result of the differential equation solution an analytical form of the current-voltage characteristic of an ionization chamber with homogeneous ionization is obtained. For the parallel-plane case comparision with experimental data is performed.

  6. Particle and Smoke Detection on ISS for Next Generation Smoke Detectors

    NASA Technical Reports Server (NTRS)

    Urban, David L.; Ruff, Gary; Yuan, Zeng-guang; Sheredy, William; Funk, Greg

    2007-01-01

    Rapid fire detection requires the ability to differentiate fire signatures from background conditions and nuisance sources. Proper design of a fire detector requires detailed knowledge of all of these signal sources so that a discriminating detector can be designed. Owing to the absence of microgravity smoke data, all current spacecraft smoke detectors were designed based upon normal-g conditions. The removal of buoyancy reduces the velocities in the high temperature zones in flames, increasing the residence time of smoke particles and consequently allowing longer growth time for the particles. Recent space shuttle experiments confirmed that, in some cases, increased particles sizes are seen in low-gravity and that the relative performance of the ISS (International Space Station) and space-shuttle smoke-detectors changes in low-gravity; however, sufficient particle size information to design new detectors was not obtained. To address this issue, the SAME (Smoke Aerosol Measurement Experiment) experiment is manifested to fly on the ISS in 2007. The SAME experiment will make measurements of the particle size distribution of the smoke particulate from several typical spacecraft materials providing quantitative design data for spacecraft smoke detectors. A precursor experiment (DAFT: Dust Aerosol measurement Feasibility Test) flew recently on the ISS and provided the first measurement of the background smoke particulate levels on the ISS. These background levels are critical to the design of future smoke detectors. The ISS cabin was found to be a very clean environment with particulate levels substantially below the space shuttle and typical ground-based environments.

  7. Use of relativistic rise in ionization chambers for measurement of high energy heavy nuclei

    NASA Technical Reports Server (NTRS)

    Barthelmy, S. D.; Israel, M. H.; Klarmann, J.; Vogel, J. S.

    1983-01-01

    A balloon-borne instrument has been constructed to measure the energy spectra of cosmic-ray heavy nuclei in the range of about 0.3 to about 100 GeV/amu. It makes use of the relativistic rise portion of the Bethe-Bloch curve in ionization chambers for energy determination in the 10- to 100-GeV/amu interval. The instrument consists of six layers of dual-gap ionization chambers for energy determination above 10 GeV/amu. Charge is determined with a NE114 scintillator and a Pilot 425 plastic Cerenkov counter. A CO2 gas Cerenkov detector (1 atm; threshold of 30 GeV/amu) calibrates the ion chambers in the relativistic rise region. The main emphasis of the instrument is the determination of the change of the ratio of Iron (26) to the Iron secondaries (21-25) in the energy range of 10 to 100 GeV/amu. Preliminary data from a balloon flight in the fall of 1982 from Palestine, TX is presented.

  8. Drift chamber vertex detectors for SLC/LEP

    SciTech Connect

    Hayes, K.G.

    1987-03-01

    The short but measurable lifetimes of the b and c quarks and the tau lepton have motivated the development of high precision tracking detectors capable of providing information on the decay vertex topology of events containing these particles. This paper reviews the OPAL, L3, and MARK II experiments vertex drift chambers.

  9. An automated ionization chamber for secondary radioactivity standards.

    PubMed

    Fitzgerald, R

    2010-01-01

    I report on the operation and characterization of a new ionization chamber system, "AUTOIC", featuring a commercial digital electrometer and a commercial robotic sample changer. The relative accuracy of the electrometer was improved significantly beyond the manufacturer's specifications through an in-house calibration of the various ranges, applied via software. The measurement precision and repeatability of the system have been determined by measuring multiple samples of the same radionuclide over the span of two or three years. The linearity of the system was examined by following the decay of (99m)Tc, (99)Mo and (133)Xe sources for up to 19 half-lives and determining half-life values. All of these values agree with the accepted literature values, within their combined uncertainties.

  10. A multiwire ionization chamber readout circuit using current mirrors

    NASA Astrophysics Data System (ADS)

    Rawnsley, W. R.; Smith, D.; Moskven, T.

    1997-01-01

    A circuit which utilizes current mirrors has been used to apply high voltage bias to the wires of a multiwire ionization chamber (MWIC) profile monitor while still allowing measurement of the beam-induced ion-electron currents collected on the wires. Bias voltages of up to 250 V have been used while wire currents over a range of 0.5 nA to 50 nA have been measured. The circuit is unipolar but can be designed for positive or negative bias. The mirrors also provide a current gain of 10, reducing the effects of transistor leakage and extending the useful range of the circuit to lower signal levels. A module containing 32 Wilson current mirrors has been constructed and is used with a MWIC monitor in TRIUMF's Parity experiment beamline.

  11. Time-zero fission-fragment detector based on low-pressure multiwire proportional chambers

    SciTech Connect

    Assamagan, Ketevi; Baker, O.; Bayatian, G.; Carlini, Roger; Danagoulian, Areg; Eden, Thomas; Egiyan, Kim; Ent, Rolf; Fenker, Howard; Gan, Liping; Gasparian, Ashot; Grigoryan, Hovhannes; Greenwood, Z; Gueye, Paul; Hashimoto, Osamu; Johnston, Kathleen; Keppel, Cynthia; Knyazian, S.; Majewski, Stanislaw; Magaryan, A; Margarian, Yu.; Marikyan, Gagik; Martoff, Charles; Mkrtchyan, Hamlet; PARLAKYAN, L.; Parlakyan, L.; Sato, Ikuro; Sawafta, Reyad; Simicevic, Neven; Tadevosyan, Vardan; Takahashi, Toshiyuki; Tang, Liguang; VARTANYAN, G.; Vulcan, William; Wells, Steven; Wood, Stephen

    1999-05-01

    A time-zero fission fragment (FF) detector, based on the technique of low-pressure multiwire proportional chambers (LPMWPC), has been designed and constructed for the heavy hypernuclear lifetime experiment (E95-002) at Thomas Jefferson National Accelerator Facility. Its characteristics and the method of time-zero reconstruction were investigated using fission fragments from a 252Cf spontaneous fission source. The influence of the ionization energy loss was also studied. It is shown that Heptane, Hexane, and Isobutane gases at a pressure of 1z2Torr are all suitable for such a FF detector. As desired by experiment, a timing resolution of about 200ps (FWHM) for a chamber size of 21z21cm2 was achieved.

  12. Integrated atom detector based on field ionization near carbon nanotubes

    SciTech Connect

    Gruener, B.; Jag, M.; Stibor, A.; Visanescu, G.; Haeffner, M.; Kern, D.; Guenther, A.; Fortagh, J.

    2009-12-15

    We demonstrate an atom detector based on field ionization and subsequent ion counting. We make use of field enhancement near tips of carbon nanotubes to reach extreme electrostatic field values of up to 9x10{sup 9} V/m, which ionize ground-state rubidium atoms. The detector is based on a carpet of multiwall carbon nanotubes grown on a substrate and used for field ionization, and a channel electron multiplier used for ion counting. We measure the field enhancement at the tips of carbon nanotubes by field emission of electrons. We demonstrate the operation of the field ionization detector by counting atoms from a thermal beam of a rubidium dispenser source. By measuring the ionization rate of rubidium as a function of the applied detector voltage we identify the field ionization distance, which is below a few tens of nanometers in front of nanotube tips. We deduce from the experimental data that field ionization of rubidium near nanotube tips takes place on a time scale faster than 10{sup -10} s. This property is particularly interesting for the development of fast atom detectors suitable for measuring correlations in ultracold quantum gases. We also describe an application of the detector as partial pressure gauge.

  13. Polarity correction factor for flattening filter free photon beams in several cylindrical ionization chambers.

    PubMed

    Ogata, Toshiyuki; Uehara, Kazuyuki; Nakayama, Masao; Tsudou, Shinji; Masutani, Takashi; Okayama, Takanobu

    2016-07-01

    In this study, we aimed to compare the polarity correction factor in ionization chambers for flattening filter free (FFF) photon beams and flattening filter (FF) beams. Measurements were performed with both 6 and 10 MV FFF and FF beams. Five commercial ionization chambers were evaluated: PTW TN30013; IBA Dosimetry CC01, CC04, and CC13; and Exradin A12S. Except for the CC01 ionization chamber, the other four chambers showed less than a 0.3 % difference in the polarity effect between the FFF and the FF beams. The CC01 chamber showed a strong field-size-dependence, unlike the other chambers. The polarity effect for all chambers with FFF beams did not change with the dose rate. Except in the case of the CC01 chamber, the difference in the polarity effect between FFF and FF beams was not significant.

  14. Evaluation of Gas-filled Ionization Chamber Method for Radon Measurement at Two Reference Facilities

    SciTech Connect

    Ishikawa, Tetsuo; Tokonami, Shinji; Kobayashi, Yosuke; Sorimachi, Atsuyuki; Yatabe, Yoshinori; Miyahara, Nobuyuki

    2008-08-07

    For quality assurance, gas-filled ionization chamber method was tested at two reference facilities for radon calibration: EML (USA) and PTB (Germany). Consequently, the radon concentrations estimated by the ionization chamber method were in good agreement with the reference radon concentrations provided by EML as well as PTB.

  15. Investigation of electric field distribution on FAC-IR-300 ionization chamber

    NASA Astrophysics Data System (ADS)

    Mohammadi, S. M.; Tavakoli-Anbaran, H.; Zeinali, H. Z.

    2016-07-01

    One of the important parameters for establishing charge particle equilibrium (CPE) conditions of free-air ionization chamber is an electric field distribution. In this paper, electric field distribution inside the ionization chamber was investigated by finite element method. For this purpose, the effects of adding guard plate and guard strips on the electric field distribution in the ionization chamber were studied. it is necessary to apply a lead box around the ionization chamber body to avoid of scattered radiation effects on the ionization chamber operation, but the lead box changes the electric field distribution. In the following, the effect of lead box on the electric field distribution was studied. Finally, electric field distribution factor (kfield) was calculated by the simulation. The results of the simulation showed that presence of the guard plate and guard strips, and applying a suitable potential to lead box, a convergence of kfield to 1 was achieved.

  16. Correlates of reported smoke detector usage in an inner-city population: participants in a smoke detector give-away program.

    PubMed Central

    Shaw, K N; McCormick, M C; Kustra, S L; Ruddy, R M; Casey, R D

    1988-01-01

    As part of a smoke detector give-away program, 388 adults were surveyed to characterize smoke detector ownership in a low-income population and to identify those who would acquire a free smoke detector following their child's visit to the hospital. Factors associated with smoke detector ownership included higher education, home ownership (vs public housing), knowledge of the city smoke detector law, and the practice of other injury prevention measures. Regardless of ownership, the great majority of parents (82 per cent) acquired a free smoke detector, but those previously without a smoke detector were more likely to do so. These characteristics of smoke detector usage and acquisition should be considered in targeting future intervention strategies. PMID:3369594

  17. Acoustic Sensor Design for Dark Matter Bubble Chamber Detectors.

    PubMed

    Felis, Ivan; Martínez-Mora, Juan Antonio; Ardid, Miguel

    2016-06-10

    Dark matter bubble chamber detectors use piezoelectric sensors in order to detect and discriminate the acoustic signals emitted by the bubbles grown within the superheated fluid from a nuclear recoil produced by a particle interaction. These sensors are attached to the outside walls of the vessel containing the fluid. The acoustic discrimination depends strongly on the properties of the sensor attached to the outer wall of the vessel that has to meet the requirements of radiopurity and size. With the aim of optimizing the sensor system, a test bench for the characterization of the sensors has been developed. The sensor response for different piezoelectric materials, geometries, matching layers, and backing layers have been measured and contrasted with FEM simulations and analytical models. The results of these studies lead us to have a design criterion for the construction of specific sensors for the next generation of dark matter bubble chamber detectors (250 L).

  18. Acoustic Sensor Design for Dark Matter Bubble Chamber Detectors

    PubMed Central

    Felis, Ivan; Martínez-Mora, Juan Antonio; Ardid, Miguel

    2016-01-01

    Dark matter bubble chamber detectors use piezoelectric sensors in order to detect and discriminate the acoustic signals emitted by the bubbles grown within the superheated fluid from a nuclear recoil produced by a particle interaction. These sensors are attached to the outside walls of the vessel containing the fluid. The acoustic discrimination depends strongly on the properties of the sensor attached to the outer wall of the vessel that has to meet the requirements of radiopurity and size. With the aim of optimizing the sensor system, a test bench for the characterization of the sensors has been developed. The sensor response for different piezoelectric materials, geometries, matching layers, and backing layers have been measured and contrasted with FEM simulations and analytical models. The results of these studies lead us to have a design criterion for the construction of specific sensors for the next generation of dark matter bubble chamber detectors (250 L). PMID:27294937

  19. Performance of the Time Expansion Chamber / Transition Radiation Detector in PHENIX Experiment at RHIC

    NASA Astrophysics Data System (ADS)

    Luiz Silva, Cesar

    2004-10-01

    The Time Expansion Chamber / Transition Radiation Detector (TEC/TRD) in the PHENIX Experiment at RHIC measures ionization losses (dE/dX) and transition radiation from charged particles produced by beam collisions. It is designed to perform tracking and identification for charged particles on very high particle multiplicity environment. The TEC/TRD consists of 24 wire chambers readout on both sides filled with recycled Xe-based gas mixture. This wire chamber configuration, besides providing measurements of ionization losses for charged particles, can absorb X-Ray photons generated by transition radiation from incident particles with γ>1000 crossing fiber radiators placed at the entrance of the chambers. This allows TEC/TRD to distinguish electrons from the huge pion signal produced over a broad momentum range (1GeV/c

  20. Higher-Sensitivity Ionization Trace-Species Detector

    NASA Technical Reports Server (NTRS)

    Boumsellek, Said; Chutjian, Ara

    1995-01-01

    Electron source and electron optics of reversal electron-attachment detector modified to increase sensitivity. Original version described in "High-Sensitivity Ionization Trace-Species Detector" (NPO-17596). Used to detect molecules of particular chemical species of interest (e.g., narcotics, explosives, or organic wastes) present in air at low concentrations, and known to attach extremely low-energy electrons. Apparatus does this by ionizing molecules from sampled atmosphere, then detecting ions of species of interest. Detector features indirectly heated spherical cathode and redesigned electron optics, together, deliver more electrons at low kinetic energy to reversal plane, R. Greater electron current generates more ions for detection.

  1. Monte Carlo calculation of energy deposition in ionization chambers for tritium measurements

    NASA Astrophysics Data System (ADS)

    Zhilin, Chen; Shuming, Peng; Dan, Meng; Yuehong, He; Heyi, Wang

    2014-10-01

    Energy deposition in ionization chambers for tritium measurements has been theoretically studied using Monte Carlo code MCNP 5. The influence of many factors, including carrier gas, chamber size, wall materials and gas pressure, has been evaluated in the simulations. It is found that β rays emitted by tritium deposit much more energy into chambers flowing through with argon than with deuterium in them, as much as 2.7 times higher at pressure 100 Pa. As chamber size gets smaller, energy deposition decreases sharply. For an ionization chamber of 1 mL, β rays deposit less than 1% of their energy at pressure 100 Pa and only 84% even if gas pressure is as high as 100 kPa. It also indicates that gold plated ionization chamber results in the highest deposition ratio while aluminum one leads to the lowest. In addition, simulations were validated by comparison with experimental data. Results show that simulations agree well with experimental data.

  2. Ionization Collection in Detectors of the Cryogenic Dark Matter Search

    SciTech Connect

    Phipps, Arran T.J.

    2016-01-01

    Determining the composition of dark matter is at the forefront of modern scientific research. There is compelling evidence for the existence of vast quantities of dark matter throughout the universe, however it has so-far eluded all direct detection efforts and its identity remains a mystery. Weakly interacting massive particles (WIMPs) are a favored dark matter candidate and have been the primary focus of direct detection for several decades. The Cryogenic Dark Matter Search (CDMS) has developed the Z-dependent Ionization and Phonon (ZIP) detector to search for such particles. Typically made from germanium, these detectors are capable of distinguishing between electromagnetic background and a putative WIMP signal through the simultaneous measurement of ionization and phonons produced by scattering events. CDMS has operated several arrays of these detectors at the Soudan Underground Laboratory (Soudan, MN, USA) resulting in many competitive (often world-leading) WIMP exclusion limits. This dissertation focuses on ionization collection in these detectors under the sub-Kelvin, low electric field, and high crystal purity conditions unique to CDMS. The design and performance of a fully cryogenic HEMT-based amplifier capable of achieving the SuperCDMS SNOLAB ionization energy resolution goal of 100 eVee is presented. The experimental apparatus which has been used to record electron and hole properties under CDMS conditions is described. Measurements of charge transport, trapping, and impact ionization as a function of electric field in two CDMS detectors are shown, and the ionization collection efficiency is determined. The data is used to predict the error in the nuclear recoil energy scale under both CDMSlite and iZIP operating modes. A two species, two state model is developed to describe how ionization collection and space charge generation in CDMS detectors are controlled by the presence of “overcharged” D- donor and A+ acceptor impurity states. The thermal

  3. Optimization of signal extraction and front-end design in a fast,multigap ionization chamber

    SciTech Connect

    Datte, P.S.; Manfredi, P.F.; Millaud, J.E.; Placidi, M.; Ratti,L.; Speziali, V.; Traversi, G.; Turner, W.C.

    2001-11-05

    This paper discusses the criteria that have been adopted tooptimize the signal processing in a shower detector to be employed as LHCbeam luminosity monitor. The original aspect ofthis instrument is itsablility to operate on a bunch-by-bunch basis. This means that it mustperform accurate charge measurements at a repetition rate of 40 MHz. Thedetector must withstand an integrated dose of 100 Grad, that is, two tothree orders of magnitude beyond those expected in the experiments. Tomeet the above requirements, an ionization chamber consisting of severalgaps of thickness 0.5 mm, filled with a gas that is expected to beradiation resistant, has been designed. Crucial in the development of thesystem is the signal processing, as the electronic noise may set thedominant limitation to the accuracy of the measurement. This is relatedto two aspects. One is the short time available for the chargemeasurement. The second one is the presence of a few meter cable betweenthe detector and the preamplifier, as this must be located out of theregion of highest radiation field. Therefore the optimization of thesignal-to-noise ratio requires that the best configuration of the chambergaps be determined under the constraint of the presence of a cable ofnon-negligible length between detector and preamplifier. The remoteplacement of the amplifying electronics will require that the front-endelectronics be radiation hard although to a lesser extent than thedetector.

  4. The control system of the multi-strip ionization chamber for the HIMM

    NASA Astrophysics Data System (ADS)

    Li, Min; Yuan, Y. J.; Mao, R. S.; Xu, Z. G.; Li, Peng; Zhao, T. C.; Zhao, Z. L.; Zhang, Nong

    2015-03-01

    Heavy Ion Medical Machine (HIMM) is a carbon ion cancer treatment facility which is being built by the Institute of Modern Physics (IMP) in China. In this facility, transverse profile and intensity of the beam at the treatment terminals will be measured by the multi-strip ionization chamber. In order to fulfill the requirement of the beam position feedback to accomplish the beam automatic commissioning, less than 1 ms reaction time of the Data Acquisition (DAQ) of this detector must be achieved. Therefore, the control system and software framework for DAQ have been redesigned and developed with National Instruments Compact Reconfigurable Input/Output (CompactRIO) instead of PXI 6133. The software is Labview-based and developed following the producer-consumer pattern with message mechanism and queue technology. The newly designed control system has been tested with carbon beam at the Heavy Ion Research Facility at Lanzhou-Cooler Storage Ring (HIRFL-CSR) and it has provided one single beam profile measurement in less than 1 ms with 1 mm beam position resolution. The fast reaction time and high precision data processing during the beam test have verified the usability and maintainability of the software framework. Furthermore, such software architecture is easy-fitting to applications with different detectors such as wire scanner detector.

  5. Two-dimensional and quasi-three-dimensional dosimetry of hadron and photon beams with the Magic Cube and the Pixel Ionization Chamber.

    PubMed

    Cirio, R; Garelli, E; Schulte, R; Amerio, S; Boriano, A; Bourhaleb, F; Coutrakon, G; Donetti, M; Giordanengo, S; Koss, P; Madon, E; Marchetto, F; Nastasi, U; Peroni, C; Santuari, D; Sardo, A; Scielzo, G; Stasi, M; Trevisiol, E

    2004-08-21

    Two detectors for fast two-dimensional (2D) and quasi-three-dimensional (quasi-3D) verification of the dose delivered by radiotherapy beams have been developed at University and Istituto Nazionale di Fisica Nucleare (INFN) of Torino. The Magic Cube is a stack of strip-segmented ionization chambers interleaved with water-equivalent slabs. The parallel plate ionization chambers have a sensitive area of 24 x 24 cm2, and consist of 0.375 cm wide and 24 cm long strips. There are a total of 64 strips per chamber. The Magic Cube has been tested with the clinical proton beam at Loma Linda University Medical Centre (LLUMC), and was shown to be capable of fast and precise quasi-3D dose verification. The Pixel Ionization Chamber (PXC) is a detector with pixel anode segmentation. It is a 32 x 32 matrix of 1024 cylindrical ionization cells arranged in a square 24 x 24 cm2 area. Each cell has 0.4 cm diameter and 0.55 cm height, at a pitch of 0.75 cm separates the centre of adjacent cells. The sensitive volume of each single ionization cell is 0.07 cm3. The detectors are read out using custom designed front-end microelectronics and a personal computer-based data acquisition system. The PXC has been used to verify dynamic intensity-modulated radiotherapy for head-and-neck and breast cancers.

  6. Two-dimensional and quasi-three-dimensional dosimetry of hadron and photon beams with the Magic Cube and the Pixel Ionization Chamber

    NASA Astrophysics Data System (ADS)

    Cirio, R.; Garelli, E.; Schulte, R.; Amerio, S.; Boriano, A.; Bourhaleb, F.; Coutrakon, G.; Donetti, M.; Giordanengo, S.; Koss, P.; Madon, E.; Marchetto, F.; Nastasi, U.; Peroni, C.; Santuari, D.; Sardo, A.; Scielzo, G.; Stasi, M.; Trevisiol, E.

    2004-08-01

    Two detectors for fast two-dimensional (2D) and quasi-three-dimensional (quasi-3D) verification of the dose delivered by radiotherapy beams have been developed at University and Istituto Nazionale di Fisica Nucleare (INFN) of Torino. The Magic Cube is a stack of strip-segmented ionization chambers interleaved with water-equivalent slabs. The parallel plate ionization chambers have a sensitive area of 24 × 24 cm2, and consist of 0.375 cm wide and 24 cm long strips. There are a total of 64 strips per chamber. The Magic Cube has been tested with the clinical proton beam at Loma Linda University Medical Centre (LLUMC), and was shown to be capable of fast and precise quasi-3D dose verification. The Pixel Ionization Chamber (PXC) is a detector with pixel anode segmentation. It is a 32 × 32 matrix of 1024 cylindrical ionization cells arranged in a square 24 × 24 cm2 area. Each cell has 0.4 cm diameter and 0.55 cm height, at a pitch of 0.75 cm separates the centre of adjacent cells. The sensitive volume of each single ionization cell is 0.07 cm3. The detectors are read out using custom designed front-end microelectronics and a personal computer-based data acquisition system. The PXC has been used to verify dynamic intensity-modulated radiotherapy for head-and-neck and breast cancers.

  7. D-IMRT verification with a 2D pixel ionization chamber: dosimetric and clinical results in head and neck cancer.

    PubMed

    Stasi, M; Giordanengo, S; Cirio, R; Boriano, A; Bourhaleb, F; Cornelius, I; Donetti, M; Garelli, E; Gomola, I; Marchetto, F; Porzio, M; Sanz Freire, C J; Sardo, A; Peroni, C

    2005-10-07

    Dynamic intensity-modulated radiotherapy (D-IMRT) using the sliding-window technique is currently applied for selected treatments of head and neck cancer at Institute for Cancer Research and Treatment of Candiolo (Turin, Italy). In the present work, a PiXel-segmented ionization Chamber (PXC) has been used for the verification of 19 fields used for four different head and neck cancers. The device consists of a 32x32 matrix of 1024 parallel-plate ionization chambers arranged in a square of 24x24 cm2 area. Each chamber has 0.4 cm diameter and 0.55 cm height; a distance of 0.75 cm separates the centre of adjacent chambers. The sensitive volume of each single ionization chamber is 0.07 cm3. Each of the 1024 independent ionization chambers is read out with a custom microelectronics chip.The output factors in water obtained with the PXC at a depth of 10 cm were compared to other detectors and the maximum difference was 1.9% for field sizes down to 3x3 cm2. Beam profiles for different field dimensions were measured with the PXC and two other types of ionization chambers; the maximum distance to agreement (DTA) in the 20-80% penumbra region of a 3x3 cm2 field was 0.09 cm. The leaf speed of the multileaf collimator was varied between 0.07 and 2 cm s-1 and the detector response was constant to better than 0.6%. The behaviour of the PXC was measured while varying the dose rate between 0.21 and 1.21 Gy min-1; the mean difference was 0.50% and the maximum difference was 0.96%. Using fields obtained with an enhanced dynamic wedge and a staircase-like (step) IMRT field, the PXC has been tested for simple 1D modulated beams; comparison with film gave a maximum DTA of 0.12 cm. The PXC was then used to check four different IMRT plans for head and neck cancer treatment: cervical chordoma, parotid, ethmoid and skull base. In the comparison of the PXC versus film and PXC versus treatment planning system, the number of pixels with gamma parameter

  8. D-IMRT verification with a 2D pixel ionization chamber: dosimetric and clinical results in head and neck cancer

    NASA Astrophysics Data System (ADS)

    Stasi, M.; Giordanengo, S.; Cirio, R.; Boriano, A.; Bourhaleb, F.; Cornelius, I.; Donetti, M.; Garelli, E.; Gomola, I.; Marchetto, F.; Porzio, M.; Sanz Freire, C. J.; Sardo, A.; Peroni, C.

    2005-10-01

    Dynamic intensity-modulated radiotherapy (D-IMRT) using the sliding-window technique is currently applied for selected treatments of head and neck cancer at Institute for Cancer Research and Treatment of Candiolo (Turin, Italy). In the present work, a PiXel-segmented ionization Chamber (PXC) has been used for the verification of 19 fields used for four different head and neck cancers. The device consists of a 32 × 32 matrix of 1024 parallel-plate ionization chambers arranged in a square of 24 × 24 cm2 area. Each chamber has 0.4 cm diameter and 0.55 cm height; a distance of 0.75 cm separates the centre of adjacent chambers. The sensitive volume of each single ionization chamber is 0.07 cm3. Each of the 1024 independent ionization chambers is read out with a custom microelectronics chip. The output factors in water obtained with the PXC at a depth of 10 cm were compared to other detectors and the maximum difference was 1.9% for field sizes down to 3 × 3 cm2. Beam profiles for different field dimensions were measured with the PXC and two other types of ionization chambers; the maximum distance to agreement (DTA) in the 20-80% penumbra region of a 3 × 3 cm2 field was 0.09 cm. The leaf speed of the multileaf collimator was varied between 0.07 and 2 cm s-1 and the detector response was constant to better than 0.6%. The behaviour of the PXC was measured while varying the dose rate between 0.21 and 1.21 Gy min-1; the mean difference was 0.50% and the maximum difference was 0.96%. Using fields obtained with an enhanced dynamic wedge and a staircase-like (step) IMRT field, the PXC has been tested for simple 1D modulated beams; comparison with film gave a maximum DTA of 0.12 cm. The PXC was then used to check four different IMRT plans for head and neck cancer treatment: cervical chordoma, parotid, ethmoid and skull base. In the comparison of the PXC versus film and PXC versus treatment planning system, the number of pixels with γ parameter <=1 was 97.7% and 97

  9. Use of a novel two-dimensional ionization chamber array for pencil beam scanning proton therapy beam quality assurance.

    PubMed

    Lin, Liyong; Kang, Minglei; Solberg, Timothy D; Mertens, Thierry; Baeumer, Christian; Ainsley, Christopher G; McDonough, James E

    2015-05-08

    The need to accurately and efficiently verify both output and dose profiles creates significant challenges in quality assurance of pencil beam scanning (PBS) proton delivery. A system for PBS QA has been developed that combines a new two-dimensional ionization chamber array in a waterproof housing that is scanned in a water phantom. The MatriXX PT has the same detector array arrangement as the standard MatriXX(Evolution) but utilizes a smaller 2 mm plate spacing instead of 5mm. Because the bias voltage of the MatriXX PT and Evolution cannot be changed, PPC40 and FC65-G ionization chambers were used to assess recombination effects. The PPC40 is a parallel plate chamber with an electrode spacing of 2mm, while the FC65-G is a Farmer chamber FC65-G with an electrode spacing of 2.8 mm. Three bias voltages (500, 200, and 100 V) were used for both detectors to determine which radiation type (continuous, pulse or pulse-scanned beam) could closely estimate Pion from the ratios of charges collected. In comparison with the MatriXX(Evolution), a significant improvement in measurement of absolute dose with the MatriXX PT was observed. While dose uncertainty of the MatriXX(Evolution) can be up to 4%, it is < 1% for the MatriXX PT. Therefore the MatriXX(Evolution) should not be used for QA of PBS for conditions in which ion recombination is not negligible. Farmer chambers should be used with caution for measuring the absolute dose of PBS beams, as the uncertainty of Pion can be > 1%; chambers with an electrode spacing of 2 mm or smaller are recommended.

  10. Use of a novel two-dimensional ionization chamber array for pencil beam scanning proton therapy beam quality assurance.

    PubMed

    Lin, Liyong; Kang, Minglei; Solberg, Timothy D; Mertens, Thierry; Baumer, Christian; Ainsley, Christopher G; McDonough, James E

    2015-05-01

    The need to accurately and efficiently verify both output and dose profiles creates significant challenges in quality assurance of pencil beam scanning (PBS) proton delivery. A system for PBS QA has been developed that combines a new two-dimensional ionization chamber array in a waterproof housing that is scanned in a water phantom. The MatriXX PT has the same detector array arrangement as the standard MatriXXEvolution but utilizes a smaller 2 mm plate spacing instead of 5 mm. Because the bias voltage of the MatriXX PT and Evolution cannot be changed, PPC40 and FC65-G ionization chambers were used to assess recombination effects. The PPC40 is a parallel plate chamber with an electrode spacing of 2 mm, while the FC65-G is a Farmer chamber FC65-G with an electrode spacing of 2.8 mm. Three bias voltages (500, 200, and 100 V) were used for both detectors to determine which radiation type (continuous, pulse or pulse-scanned beam) could closely estimate Pion from the ratios of charges collected. In comparison with the MatriXXEvolution, a significant improvement in measurement of absolute dose with the MatriXX PT was observed. While dose uncertainty of the MatriXXEvolution can be up to 4%, it is <1% for the MatriXX PT. Therefore the MatriXXEvolution should not be used for QA of PBS for conditions in which ion recombination is not negligible. Farmer chambers should be used with caution for measuring the absolute dose of PBS beams, as the uncertainty of Pion can be <1%; chambers with an electrode spacing of 2 mm or smaller are recommended. PACS number: 87.53.Qc.

  11. Construction of an ionization chamber for the measurement of dose of low energy x-rays

    SciTech Connect

    Perez, Y. B. Alcantara; Jimenez, F. J. Ramirez

    2008-08-11

    We designed and constructed the prototype of an ionization chamber to measure the dose of an X-ray tube with Molybdenum anode. This X-ray tube is located in the Physics department at CINVESTAV and is used for medical physics purposes in the imaging area. The ionization chamber is designed to measure doses on biological samples exposed to X-rays and will be applied in radiation protection studies.

  12. Pencil beam proton radiography using a multilayer ionization chamber

    NASA Astrophysics Data System (ADS)

    Farace, Paolo; Righetto, Roberto; Meijers, Arturs

    2016-06-01

    A pencil beam proton radiography (PR) method, using a commercial multilayer ionization chamber (MLIC) integrated with a treatment planning system (TPS) was developed. A Giraffe (IBA Dosimetry) MLIC (±0.5 mm accuracy) was used to obtain pencil beam PR by delivering spots uniformly positioned at a 5.0 mm distance in a 9  ×  9 square of spots. PRs of an electron-density (with tissue-equivalent inserts) phantom and a head phantom were acquired. The integral depth dose (IDD) curves of the delivered spots were computed by the TPS in a volume of water simulating the MLIC, and virtually added to the CT at the exit side of the phantoms. For each spot, measured and calculated IDD were overlapped in order to compute a map of range errors. On the head-phantom, the maximum dose from PR acquisition was estimated. Additionally, on the head phantom the impact on the range errors map was estimated in case of a 1 mm position misalignment. In the electron-density phantom, range errors were within 1 mm in the soft-tissue rods, but greater in the dense-rod. In the head-phantom the range errors were  -0.9  ±  2.7 mm on the whole map and within 1 mm in the brain area. On both phantoms greater errors were observed at inhomogeneity interfaces, due to sensitivity to small misalignment, and inaccurate TPS dose computation. The effect of the 1 mm misalignment was clearly visible on the range error map and produced an increased spread of range errors (-1.0  ±  3.8 mm on the whole map). The dose to the patient for such PR acquisitions would be acceptable as the maximum dose to the head phantom was  <2cGyE. By the described 2D method, allowing to discriminate misalignments, range verification can be performed in selected areas to implement an in vivo quality assurance program.

  13. Performances of a VLSI wide dynamic range current-to-frequency converter for strip ionization chambers

    NASA Astrophysics Data System (ADS)

    Bonazzola, G. C.; Cirio, R.; Donetti, M.; Marchetto, F.; Mazza, G.; Peroni, C.; Zampieri, A.

    1998-02-01

    In this paper we report on the design and test of a 14-channel VLSI chip to perform the current to frequency conversion for parallel plate strip ionization chambers. The chambers measure the intensity and the geometrical characteristics of a therapeutical beam.

  14. A pixel ionization chamber used as beam monitor at the Institut Curie—Centre de Protontherapie de Orsay (CPO)

    NASA Astrophysics Data System (ADS)

    La Rosa, A.; Garella, M. A.; Bourhaleb, F.; Cirio, R.; Donetti, M.; Giordanengo, S.; Givehchi, N.; Marchetto, F.; Martin, F.; Meyroneinc, S.; Peroni, C.; Pittà, G.

    2006-09-01

    The Dipartimento di Fisica Sperimentale and the Istituto Nazionale di Fisica Nucleare (INFN), Torino, in collaboration with the Institut Curie—Centre de Protontherapie de Orsay (CPO), have developed and built a pixel parallel plate ionization chamber to be used as monitor for the proton therapy beam line at the Institut Curie—CPO (Orsay, France). The sensitive area of the detector is (160×160) mm 2, with the anode segmented in 1024 square pixels arranged in a 32×32 matrix; the area of each pixel is (5×5) mm 2. The detector has been placed on the beam line just upstream of the last collimator to monitor the beam shape and to measure the stability and reproducibility of the delivery system. In this paper, we present a detailed description of the detector and the results of a set of preliminary tests.

  15. Preliminary results with a strip ionization chamber used as beam monitor for hadrontherapy treatments

    NASA Astrophysics Data System (ADS)

    Boriano, A.; Bourhaleb, F.; Cirio, R.; Cirrone, G. A. P.; Cuttone, G.; Donetti, M.; Garelli, E.; Giordanengo, S.; Luparia, A.; Marchette, F.; Peroni, C.; Raffaele, L.; Sabini, M. G.; Valastro, L.

    2006-01-01

    Preliminary results are presented from a test of a parallel plate ionization chamber with the anode segmented in strips (MOPI) to be used as a beam monitor for therapeutical treatments on the 62 MeV proton beam line of the INFN-LNS Superconducting Cyclotron. Ocular pathologies have been treated at the Catana facility since March 2002. The detector, placed downstream of the patient collimator, will allow the measurement of the relevant beam diagnostic parameters during treatment such as integrated beam fluence, for dose determination; the beam baricentre, width and asymmetry will be obtained from the fluence profile sampled with a resolution of about 100 Urn at a rate up to 1 kHz with no dead time. In this test, carried out at LNS, the detector has been exposed to different beam shapes and the integrated fluence derived by the measured beam profiles has been compared with that obtained with other dosimeters normally used for treatment. The skewness of the beam profile has been measured and shown to be suitable to on-line check variations of the beam shape.

  16. The PRIME Lab gas ionization detector

    NASA Astrophysics Data System (ADS)

    Knies, David L.; Elmore, David

    1994-06-01

    A gas ionization detection system was built for optimal identification of AMS radionuclides, in particular 10Be and 36Cl. For 36Cl, a combination of 1) the difference in arrival times for electrons at two anode plates and 2) a novel split anode plate has led to a reduction in misidentified 36S. A peak-stabilizing routine incorporated in the data acquisition system has allowed us to run at higher counting rates. Changing to propane gas has reduced random signal amplitude shifts.

  17. Particle identification in a LKr ionization chamber by multiple induced current measurements using the shape analysis of the signal

    NASA Astrophysics Data System (ADS)

    Diaferia, R.; Lanni, F.; Maggi, B.; Palombo, F.; Sala, A.; Cantoni, P.; Frabetti, P. L.; Stagni, L.

    1996-01-01

    Charged particle (π/K) separation in the momentum range 0.5-0.7 GeV/c using a new method of shape analysis of the signal from a liquid krypton ionization chamber has been studied experimentally. The detector has been exposed to pions and protons at the T11 test beam at CERN PS. The shape of preamplifier output signal has been recorded by a waveform digitizer and differentiated to obtain multiple measurements of induced current inside a 2 cm gap. Results on particle separation are presented and compared with a Monte Carlo simulation.

  18. Particle identification in a LKr ionization chamber by multiple induced current measurements using the shape analysis of the signal

    NASA Astrophysics Data System (ADS)

    Cantoni, P.; Frabetti, P. L.; Stagni, L.; Diaferia, R.; Lanni, F.; Maggi, B.; Palombo, F.; Sala, A.; Manfredi, P. F.; Re, V.; Speziali, V.

    1995-02-01

    Charged particle ( {π}/{K}) separation in the momentum range 0.5-0.7 GeV/ c using a new method of shape analysis of the signal from a liquid krypton ionization chamber has been studied experimentally. The detector has been exposed to the T11 test beam at CERN PS. The shape of the preamplifier output signal has been recorded by a waveform digitizer and differentiated to obtain multiple measurements of induced current inside a 2 cm gap. Results on particle separation are presented.

  19. Investigation of the initial and volume recombination losses in gamma versatile cylindrical ionization chamber VGIC developed for gamma ray dosimetry

    SciTech Connect

    Fares, M.; Mameri, S.; Abdlani, I.; Negara, K.

    2015-07-01

    A versatile Gamma ionization chambers are used for flow control in systems with gamma nuclear reactors and reprocessing plants in and monitoring atmosphere around these facilities, this in order to protect staff. In the Laboratory Detection and Measures (LDM) Division for Study and Development of Nuclear Instrumentation (DSDNI) of CRNB, we designed, developed and characterized a versatile gamma ionization chamber (VGIC) to study experimentally its characteristics according to the geometry of the electrodes, the volume and pressure of the filler gas for the design of a gamma sealed chamber. The tests were conducted under the IEC (International Electro-technical Commission). In this paper, we present the results obtained in the various nuclear tests for characterization and calibration that we have made on the ionization chamber gamma VGIC prototype developed at our Department. To do this, three irradiators were operated at the Laboratory Calibration (SSDL) of the Department of Medical Physics Nuclear Research Center of Algiers (CRNA). Irradiator intensive gamma ({sup 60}Co: 1.25 MeV), one medium intensity gamma ({sup 137}Cs: 0.662 MeV) and 3rd low intensity ({sup 60}Co). Saturation curves and linearity were identified and the operating range and the sensitivity of the chamber have been deducted. The (I,V) characteristics of the chamber filled, with argon gas at 3 bar (0.3 M pa) pressure, for gamma ray irradiator sources were studied. To do so, the chamber was irradiated with gamma rays using different numbers of gamma sources (i.e. Up to 5). The plateau region is reached above 200 V and the detector operating voltage is found to be 600 V. It is observed that in the plateau region the slope is constant with an increase in the exposure rate. The (1/I, 1/V) and (I, l/V{sup 2}) characteristic curves reveal the presence of the initial and volume recombination losses. The volume recombination losses are found to be smaller than the initial recombination losses. Finally

  20. Boronated Scintillator Detector for Use in Space with Ionization Calorimeters

    NASA Astrophysics Data System (ADS)

    Britvich, G. I.; Chernichenko, S. K.; Demichev, M. A.; Gnezdilov, I. I.; Mukhin, V. I.; Soukhih, A. V.

    2016-02-01

    Boronated Scintillator Detector (BSD) for use in space with ionization calorimeters was suggested. BSD improved e/h showers separation, which are initiated in the ionization calorimeter in interaction it with high energy particles. Improve the rejection is based on the hadron-induced showers tend to be accompanied by significantly more neutron activity than electromagnetic showers. The detector is composed of natural boron-loaded (5%) castable plastic scintillation plates. To collect light using wavelength-shifting (WLS) fibers. The experiment showed that the photoelectron yield is ∼ 40 ph.el./MeV with using PMT EMI 9954KB. Simulation on GEANT4 was obtained neutron detection efficiency. The simulation was conducted in the assumption that neutrons have the spectrum 252Cf and fall plane-parallel on the entry surface of the detector.

  1. Polarity and ion recombination corrections in continuous and pulsed beams for ionization chambers with high Z chamber walls.

    PubMed

    Aldosary, Ghada; Safigholi, Habib; Song, William; Seuntjens, Jan; Sarfehnia, Arman

    2017-03-01

    In this work, the response of Farmer-type ionization chambers fitted with high atomic number (Z) walls is studied, and results of the effects of such walls on polarity and ion recombination correction factors in both continuous and pulsed beams are presented. Measurements were made in a continuous Co-60 beam and a pulsed 6MV linac beam using an Exradin-A12 ionization chamber fitted with the manufacturer's C-552 plastic wall, as well as geometrically identical walls made from aluminum, copper and molybdenum. The bias voltage was changed between 10values (range: +50 to +560V). Ion recombination was determined from Jaffé plots and by using the "two-voltage technique". The saturation charge measured with each chamber wall was extrapolated from Jaffé plots. Additionally, the effect of different wall materials on chamber response was studied using MCNP simulations. Results showed that the polarity correction factor is not significantly affected by changes in chamber wall material (within 0.1%). Furthermore, although the saturation charges greatly vary with each chamber wall material, and charge multiplication increases for higher atomic number wall materials, the standard methods of calculating ion recombination yielded results that differed by only 0.2%. Therefore, polarity and ion recombination correction factors are not greatly affected by the chamber wall material. The experimental saturation charges for all the different wall materials agreed well within the uncertainty with MCNP simulations. The breakdown of the linear relationship in Jaffé plots that was previously reported to exist for conventional chamber walls was also observed with the different wall materials.

  2. Variable pressure ionization detector for gas chromatography

    DOEpatents

    Buchanan, Michelle V.; Wise, Marcus B.

    1988-01-01

    Method and apparatus for differentiating organic compounds based on their electron affinity. An electron capture detector cell (ECD) is operated at pressures ranging from atmospheric to less than 1 torr. Through variation of the pressure within the ECD cell, the organic compounds are induced to either capture or emit electrons. Differentiation of isomeric compounds can be obtianed when, at a given pressure, one isomer is in the emission mode and the other is in the capture mode. Output of the ECD is recorded by chromatogram. The invention also includes a method for obtaining the zero-crossing pressure of a compound, defined as the pressure at which the competing emission and capture reactions are balanced and which may be correlated to the electron affinity of a compound.

  3. Time-resolved dosimetry using a pinpoint ionization chamber as quality assurance for IMRT and VMAT

    SciTech Connect

    Louwe, Robert J. W. Satherley, Thomas; Day, Rebecca A.; Greig, Lynne; Wendling, Markus; Monshouwer, René

    2015-04-15

    Purpose: To develop a method to verify the dose delivery in relation to the individual control points of intensity modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) using an ionization chamber. In addition to more effective problem solving during patient-specific quality assurance (QA), the aim is to eventually map out the limitations in the treatment chain and enable a targeted improvement of the treatment technique in an efficient way. Methods: Pretreatment verification was carried out for 255 treatment plans that included a broad range of treatment indications in two departments using the equipment of different vendors. In-house developed software was used to enable calculation of the dose delivery for the individual beamlets in the treatment planning system (TPS), for data acquisition, and for analysis of the data. The observed deviations were related to various delivery and measurement parameters such as gantry angle, field size, and the position of the detector with respect to the field edge to distinguish between error sources. Results: The average deviation of the integral fraction dose during pretreatment verification of the planning target volume dose was −2.1% ± 2.2% (1 SD), −1.7% ± 1.7% (1 SD), and 0.0% ± 1.3% (1 SD) for IMRT at the Radboud University Medical Center (RUMC), VMAT (RUMC), and VMAT at the Wellington Blood and Cancer Centre, respectively. Verification of the dose to organs at risk gave very similar results but was generally subject to a larger measurement uncertainty due to the position of the detector at a high dose gradient. The observed deviations could be related to limitations of the TPS beam models, attenuation of the treatment couch, as well as measurement errors. The apparent systematic error of about −2% in the average deviation of the integral fraction dose in the RUMC results could be explained by the limitations of the TPS beam model in the calculation of the beam penumbra. Conclusions: This

  4. Evaluation of a digital optical ionizing radiation particle track detector

    SciTech Connect

    Hunter, S.R.

    1987-06-01

    An ionizing radiation particle track detector is outlined which can, in principle, determine the three-dimensional spatial distribution of all the secondary electrons produced by the passage of ionizing radiation through a low-pressure (0.1 to 10 kPa) gas. The electrons in the particle track are excited by the presence of a high-frequency AC electric field, and two digital cameras image the optical radiation produced in electronic excitation collisions of the surroundings gas by the electrons. The specific requirements of the detector for neutron dosimetry and microdosimetry are outlined (i.e., operating conditions of the digital cameras, high voltage fields, gas mixtures, etc.) along with an estimate of the resolution and sensitivity achievable with this technique. The proposed detector is shown to compare favorable with other methods for obtaining the details of the track structure, particularly in the quality of the information obtainable about the particle track and the comparative simplicity and adaptability of the detector for measuring the secondary electron track structure for many forms of ionizing radiation over a wide range of energies.

  5. Miniature metastable ionization detectors for exobiology flight experiments

    NASA Technical Reports Server (NTRS)

    Woeller, F. H.

    1986-01-01

    The Metastable Ionization Detector (MID) is three orders of magnitude more sensitive than the thermal conductivity detectors used on previous flight instruments. The miniature MID provides scientists with a much smaller and highly sensitive detector for flight gas chromatographs. A miniature MID featuring an unconventional triaxial electrode configuration was developed and used routinely in the laboratory. Although much smaller and lighter than the commercial MID, its performance characteristics parallel those of the traditional design. The detector is compatible with the modulated voltage circuitry, also developed here, and thus can perform over an expanded response range of more than 7 orders magnitude. A micro volume version of a miniature MID, with an internal volume of less than 8 microliter, was recently designed is now being tested. The micro volume MID uses carrier gas flow rates of approx. 2cc/min thus eliminating the need for makeup gas when capillary columns are used.

  6. Quantification of Cigarette Smoke Particle Deposition In Vitro Using a Triplicate Quartz Crystal Microbalance Exposure Chamber

    PubMed Central

    Adamson, Jason; Thorne, David; McAughey, John; Dillon, Deborah; Meredith, Clive

    2013-01-01

    There are a variety of smoke exposure systems available to the tobacco industry and respiratory toxicology research groups, each with their own way of diluting/delivering smoke to cell cultures. Thus a simple technique to measure dose in vitro needs to be utilised. Dosimetry—assessment of dose—is a key element in linking the biological effects of smoke generated by various exposure systems. Microbalance technology is presented as a dosimetry tool and a way of measuring whole smoke dose. Described here is a new tool to quantify diluted smoke particulate deposition in vitro. The triplicate quartz crystal microbalance (QCM) chamber measured real-time deposition of smoke at a range of dilutions 1 : 5–1 : 400 (smoke : air). Mass was read in triplicate by 3 identical QCMs installed into one in vitro exposure chamber, each in the location in which a cell culture would be exposed to smoke at the air-liquid interface. This resulted in quantification of deposited particulate matter in the range 0.21–28.00 μg/cm2. Results demonstrated that the QCM could discriminate mass between dilutions and was able to give information of regional deposition where cell cultures would usually be exposed within the chamber. Our aim is to use the QCM to support the preclinical (in vitro) evaluation of tobacco products. PMID:23484139

  7. Theoretical study of energy deposition in ionization chambers for tritium measurements

    SciTech Connect

    Chen, Zhilin; Peng, Shuming; Meng, Dan; He, Yuehong; Wang, Heyi

    2013-10-15

    Energy deposition in ionization chambers has been theoretically studied for tritium measurements in gaseous form. A one-dimension model is introduced to establish the quantitative relationship between energy deposition rate and many factors, including carrier gas, gas pressure, wall material, chamber size, and gas temperature. Energy deposition rate has been calculated at pressure varying from 5 kPa to 500 kPa based on some approximations. It is found that energy deposition rate varies greatly for different parameters, especially at low gas pressure. For the same chamber, energy deposition rate in argon is much higher than in deuterium, as much as 70.7% higher at 5 kPa. Gold plated chamber gives highest energy deposition rate in the calculations while aluminum chamber results in the lowest. As chamber size gets smaller, β ray emitted by tritium will deposit less energy in the sensitive region of the chamber. For chambers flowing through with the same gas, energy deposition rate in a 10 L chamber is 23.9% higher than in a 0.05 L chamber at 5 kPa. Gas temperature also places slight influence on energy deposition rate, and 373 K will lead to 6.7% lower deposition rate than 233 K at 5 kPa. In addition, experiments have been performed to obtain energy deposition rate in a gold plated chamber, which show good accordance with theoretical calculations.

  8. Theoretical study of energy deposition in ionization chambers for tritium measurements

    NASA Astrophysics Data System (ADS)

    Chen, Zhilin; Peng, Shuming; Meng, Dan; He, Yuehong; Wang, Heyi

    2013-10-01

    Energy deposition in ionization chambers has been theoretically studied for tritium measurements in gaseous form. A one-dimension model is introduced to establish the quantitative relationship between energy deposition rate and many factors, including carrier gas, gas pressure, wall material, chamber size, and gas temperature. Energy deposition rate has been calculated at pressure varying from 5 kPa to 500 kPa based on some approximations. It is found that energy deposition rate varies greatly for different parameters, especially at low gas pressure. For the same chamber, energy deposition rate in argon is much higher than in deuterium, as much as 70.7% higher at 5 kPa. Gold plated chamber gives highest energy deposition rate in the calculations while aluminum chamber results in the lowest. As chamber size gets smaller, β ray emitted by tritium will deposit less energy in the sensitive region of the chamber. For chambers flowing through with the same gas, energy deposition rate in a 10 L chamber is 23.9% higher than in a 0.05 L chamber at 5 kPa. Gas temperature also places slight influence on energy deposition rate, and 373 K will lead to 6.7% lower deposition rate than 233 K at 5 kPa. In addition, experiments have been performed to obtain energy deposition rate in a gold plated chamber, which show good accordance with theoretical calculations.

  9. An improved leakage current compensation technique for a 4πγ ionization chamber system

    NASA Astrophysics Data System (ADS)

    Hino, Y.; Kawada, Y.; Nazaroh

    1996-02-01

    A current integration method using a small capacitor is most commonly employed for precise measurements of small currents from 4πγ ionization chambers for the secondary standardization of radionuclides. An improved technique has been developed for eliminating the possible effect due to the electrical leakage and/or current loss across the feedback capacitor used in the integration of the ionization current. This method is based upon charge integration from a fixed negative level of potential to nearly the same level of positive potential via the zero point. The validity of this method is demonstrated for some typical applications of 4πγ ionization chamber systems. This technique can contribute to the improvement of accuracy and to the extension of the intensity range of radioactive source in 4πγ ionization measurement.

  10. Determination of (137)Cs half-life with an ionization chamber.

    PubMed

    Juget, Frédéric; Nedjadi, Youcef; Buchillier, Thierry; Bochud, François; Bailat, Claude

    2016-12-01

    The half-life of (137)Cs was measured with an ionization chamber by following the decay of 5 sources over a 30 years period between 1983 and 2013. The ratio between the ionization chamber current for the cesium sources and (226)Ra source was used for the half-life calculation. The value found for the (137)Cs half-life is 10,955.2±10.7 days, where the uncertainty evaluation combines type A and B for one standard deviation.

  11. Update of NIST half-life results corrected for ionization chamber source-holder instability.

    PubMed

    Unterweger, M P; Fitzgerald, R

    2014-05-01

    As reported at the ICRM 2011, it was discovered that the source holder used for calibrations in the NIST 4πγ ionization chamber (IC) was not stable. This has affected a large number of half-life measurement results previously reported and used in compilations of nuclear data. Corrections have been made on all of the half-life data based on the assumption that the changes to the ionization chamber response were gradual. The corrections are energy dependent and therefore radionuclide specific. This presentation will review our results and present the recommended changes in half-life values and/or uncertainties.

  12. A magnetically levitated electrode ionization chamber of the noncontact measurement type

    NASA Astrophysics Data System (ADS)

    Kawaguchi, Toshiro; Yoshimura, Atsushi

    2002-04-01

    A new type of ionization chamber with levitated electrode has been developed. In this ionization chamber, an ion-collection electrode levitates in the air without getting any physical support from the insulator. The electrode is charged by an electrostatic charger without physical contact. The charge of the electrode is read out at a Faraday cage periodically at a given time interval without physical contact. Because its electrode levitates, the ionization chamber produces no background current caused by leaks or piezo current. In addition, as the charging of its electrode and the read-out of its charge are carried out without physical contact, no irregular charge or contact potential difference due to the chattering between electrode and contact point occurs. Through experiments, it was found that this ionization chamber was able to measure the γ-ray dose such as the environmental radiation with a high degree of sensitivity. The minimum detectable value of ionization current when accumulated for 1 h is about 1.3×10 -17 A.

  13. Extraction of depth-dependent perturbation factors for parallel-plate chambers in electron beams using a plastic scintillation detector

    SciTech Connect

    Lacroix, Frederic; Guillot, Mathieu; McEwen, Malcolm; Cojocaru, Claudiu; Gingras, Luc; Beddar, A. Sam; Beaulieu, Luc

    2010-08-15

    Purpose: This work presents the experimental extraction of the overall perturbation factor P{sub Q} in megavoltage electron beams for NACP-02 and Roos parallel-plate ionization chambers using a plastic scintillation detector (PSD). Methods: The authors used a single scanning PSD mounted on a high-precision scanning tank to measure depth-dose curves in 6, 12, and 18 MeV clinical electron beams. The authors also measured depth-dose curves using the NACP-02 and PTW Roos chambers. Results: The authors found that the perturbation factors for the NACP-02 and Roos chambers increased substantially with depth, especially for low-energy electron beams. The experimental results were in good agreement with the results of Monte Carlo simulations reported by other investigators. The authors also found that using an effective point of measurement (EPOM) placed inside the air cavity reduced the variation of perturbation factors with depth and that the optimal EPOM appears to be energy dependent. Conclusions: A PSD can be used to experimentally extract perturbation factors for ionization chambers. The dosimetry protocol recommendations indicating that the point of measurement be placed on the inside face of the front window appear to be incorrect for parallel-plate chambers and result in errors in the R{sub 50} of approximately 0.4 mm at 6 MeV, 1.0 mm at 12 MeV, and 1.2 mm at 18 MeV.

  14. Multiplexed electronically programmable multimode ionization detector for chromatography

    DOEpatents

    Wise, Marcus B.; Buchanan, Michelle V.

    1989-01-01

    Method and apparatus for detecting and differentiating organic compounds based on their electron affinity. An electron capture detector cell (ECD) is operated in a plurality of multiplexed electroncially programmable operating modes to alter the detector response during a single sampling cycle to acquire multiple simultaneous chromatograms corresponding to each of the different operating modes. The cell is held at a constant subatmospheric pressure while the electron collection bias voltage applied to the cell is modulated electronically to allow acquisition of multiple chromatograms for a single sample elution from a chromatograph representing three distinctly different response modes. A system is provided which automatically controls the programmed application of bias pulses at different intervals and/or amplitudes to switch the detector from an ionization mode to the electron capture mode and various degrees therebetween to provide an improved means of tuning an ECD for multimode detection and improved specificity.

  15. Multiplexed electronically programmable multimode ionization detector for chromatography

    DOEpatents

    Wise, M.B.; Buchanan, M.V.

    1988-05-19

    Method and apparatus for detecting and differentiating organic compounds based on their electron affinity. An electron capture detector cell (ECD) is operated in a plurality of multiplexed electronically programmable operating modes to alter the detector response during a single sampling cycle to acquire multiple simultaneous chromatograms corresponding to each of the different operating modes. The cell is held at a constant subatmospheric pressure while the electron collection bias voltage applied to the cell is modulated electronically to allow acquisition of multiple chromatograms for a single sample elution from a chromatograph representing three distinctly different response modes. A system is provided which automatically controls the programmed application of bias pulses at different intervals and/or amplitudes to switch the detector from an ionization mode to the electron capture mode and various degrees therebetween to provide an improved means of tuning an ECD for multimode detection and improved specificity. 6 figs.

  16. Tobacco smoke aging in the presence of ozone: A room-sized chamber study

    NASA Astrophysics Data System (ADS)

    Petrick, Lauren M.; Sleiman, Mohamad; Dubowski, Yael; Gundel, Lara A.; Destaillats, Hugo

    2011-09-01

    Exposure to tobacco pollutants that linger indoors after smoking has taken place ( thirdhand smoke, THS) can occur over extended periods and is modulated by chemical processes involving atmospheric reactive species. This study investigates the role of ozone and indoor surfaces in chemical transformations of tobacco smoke residues. Gas and particle constituents of secondhand smoke (SHS) as well as sorbed SHS on chamber internal walls and model materials (cotton, paper, and gypsum wallboard) were characterized during aging. After smoldering 10 cigarettes in a 24-m 3 room size chamber, gas-phase nicotine was rapidly removed by sorption to chamber surfaces, and subsequently re-emitted during ventilation with clean air to a level of ˜10% that during the smoking phase. During chamber ventilation in the presence of ozone (180 ppb), ozone decayed at a rate of 5.6 h -1 and coincided with a factor of 5 less nicotine sorbed to wallboard. In the presence of ozone, no gas phase nicotine was detected as a result of re-emission, and higher concentrations of nicotine oxidation products were observed than when ventilation was performed with ozone-free air. Analysis of the model surfaces showed that heterogeneous nicotine-ozone reaction was faster on paper than cotton, and both were faster than on wallboard. However, wallboard played a dominant role in ozone-initiated reaction in the chamber due to its large total geometric surface area and sink potential compared to the other substrates. This study is the first to show in a room-sized environmental chamber that the heterogeneous ozone chemistry of sorbed nicotine generates THS constituents of concern, as observed previously in bench-top studies. In addition to the main oxidation products (cotinine, myosmine and N-methyl formamide), nicotine-1-oxide was detected for the first time.

  17. Determination of relative ion chamber calibration coefficients from depth-ionization measurements in clinical electron beams

    NASA Astrophysics Data System (ADS)

    Muir, B. R.; McEwen, M. R.; Rogers, D. W. O.

    2014-10-01

    A method is presented to obtain ion chamber calibration coefficients relative to secondary standard reference chambers in electron beams using depth-ionization measurements. Results are obtained as a function of depth and average electron energy at depth in 4, 8, 12 and 18 MeV electron beams from the NRC Elekta Precise linac. The PTW Roos, Scanditronix NACP-02, PTW Advanced Markus and NE 2571 ion chambers are investigated. The challenges and limitations of the method are discussed. The proposed method produces useful data at shallow depths. At depths past the reference depth, small shifts in positioning or drifts in the incident beam energy affect the results, thereby providing a built-in test of incident electron energy drifts and/or chamber set-up. Polarity corrections for ion chambers as a function of average electron energy at depth agree with literature data. The proposed method produces results consistent with those obtained using the conventional calibration procedure while gaining much more information about the behavior of the ion chamber with similar data acquisition time. Measurement uncertainties in calibration coefficients obtained with this method are estimated to be less than 0.5%. These results open up the possibility of using depth-ionization measurements to yield chamber ratios which may be suitable for primary standards-level dissemination.

  18. SU-E-T-415: An Ionization Chamber Array with High Spatial Resolution for External Beam Radiotherapy

    SciTech Connect

    Togno, M; Wilkens, J; Menichelli, D

    2014-06-01

    Purpose: To characterize an ionization chamber array technology with high spatial resolution and high charge collection efficiency for external beam radiotherapy. Methods: The prototype under test is a linear array of air vented ionization chambers developed by IBA Dosimetry, consisting of 80 pixels with 3.5mm spatial resolution and 4mm{sup 3} sensitive volume. The detector was characterized in a plastic phantom with {sup 60} Co radiation and MV X-rays from an ELEKTA Agility LINAC (with flattened and unflattened beam qualities). Bias voltage was varied in order to evaluate charge collection efficiency. A commercial array of ionization chambers (MatriXX Evolution, IBA Dosimetry) and an amorphous silicon flat panel in direct conversion configuration were used as references. Results: Repeatability (0.4%) and stability under continuous gamma irradiation (0.3%) are very good, in spite of low active volume and sensitivity (∼200pC/Gy). Charge collection efficiency is higher than 99% already at 150V with ∼2mGy dose per pulse, leading to a ±1.1% sensitivity change with dose per pulse in the range 0.09-2mGy (covering all flattened and unflattened applications). Measured dose profiles are in agreement with MatriXX for fields larger than 2×2cm{sup 2}, in which case the linear array offers a much better characterization of the penumbra region. Down to 1×1cm{sup 2}, measured profiles are in very good agreement with the flat panel. Conclusion: The array represents a valuable tool for the characterization of treatment fields in which high spatial resolution is required, together with the dosimetric performance of air vented ionization chambers. Such a technology would be particularly valuable in association with advanced treatment modalities such as rotational radiotherapy, stereotactic treatments (even with unflattened beam qualities) and proton therapy, due to the insensitivity of the chambers on dose per pulse. In the future, a two dimensional prototype based on this

  19. Absorbed dose dependence of the correction factors for ionization chamber cable irradiation effects.

    PubMed

    Campos, L L; Caldas, L V

    1991-03-01

    A simple method was developed, for possible use by hospital physicists, to evaluate the irradiation effects on cables and connectors during large-radiation-field dosimetry with ionization chambers and to determine correction factors for the used system or geometry. This method was based on the absorbed dose dependence of the correction factor.

  20. A VLSI analog pipeline read-out for electrode segmented ionization chambers

    NASA Astrophysics Data System (ADS)

    Bonazzola, G. C.; Bouvier, S.; Cirio, R.; Donetti, M.; Figus, M.; Marchetto, F.; Peroni, C.; Pernigotti, E.; Thenard, J. M.; Zampieri, A.

    1999-05-01

    We report on the design and test of a 32-channel VLSI chip based on the analog pipeline memory concept. The charge from a strip of a ionization chamber, is stored as a function of time in a switched capacitor array. The cell reading can be done in parallel with the writing.

  1. Stability of A-150 plastic ionization chamber response over a ~30 year period

    SciTech Connect

    Kroc, Thomas K.; Lennox, Arlene J.; /Fermilab

    2007-08-01

    At the NIU Institute for Neutron Therapy at Fermilab, the clinical tissue-equivalent ionization chamber response is measured every treatment day using a cesium source that was configured to match readings obtained at the National Bureau of Standards. Daily measurements are performed in air using the air-to-tissue dose conversion factors given in AAPM Report no. 7. The measured exposure calibration factors have been tabulated and graphed as a function of time from 1978 to present. For A-150 plastic ionization chambers, these factors exhibit a sinusoidal variation with a period of approximately one year and amplitude of {+-} 1%. This variation, attributable to the hygroscopic nature of A-150 plastic, is correlated with the relative humidity of the facility, and is greater than the humidity corrections for gas described in the literature. Our data suggest that chamber calibration should be performed at least weekly to accommodate these variations.

  2. Free-air ionization chamber, FAC-IR-300, designed for medium energy X-ray dosimetry

    NASA Astrophysics Data System (ADS)

    Mohammadi, S. M.; Tavakoli-Anbaran, H.; Zeinali, H. Z.

    2017-01-01

    The primary standard for X-ray photons is based on parallel-plate free-air ionization chamber (FAC). Therefore, the Atomic Energy Organization of Iran (AEOI) is tried to design and build the free-air ionization chamber, FAC-IR-300, for low and medium energy X-ray dosimetry. The main aim of the present work is to investigate specification of the FAC-IR-300 ionization chamber and design it. FAC-IR-300 dosimeter is composed of two parallel plates, a high voltage (HV) plate and a collector plate, along with a guard electrode that surrounds the collector plate. The guard plate and the collector were separated by an air gap. For obtaining uniformity in the electric field distribution, a group of guard strips was used around the ionization chamber. These characterizations involve determining the exact dimensions of the ionization chamber by using Monte Carlo simulation and introducing correction factors.

  3. Home radon monitor modeled after the common smoke detector

    SciTech Connect

    Bolton, R.D.; Arnone, G.J.; Johnson, J.P.

    1995-02-01

    The EPA has declared that five million or so of the nation`s 80 million homes may have indoor radon levels that pose an unacceptably high risk of lung cancer to occupants. They estimate that four times as many people die from radon-induced lung cancers as from fires in the home. Therefore the EPA has recommended that all homes be tested and that action be taken to reduce the radon concentration in homes that test above the 4 pCi/L level. The push to have homeowners voluntarily test for elevated radon levels has been only marginally successful. A reliable, inexpensive, and accurate in-home radon monitor designed along the same general lines as a home smoke detector might overcome much of the public reluctance to test homes for radon. Such a Home Radon Monitor (HRM) is under development at Los Alamos National Laboratory. To be acceptable to the public, HRMs should have the following characteristics in common with smoke detectors: low cost, small size, ease of installation and use, low maintenance, and high performance. Recent advances in Long-Range Alpha Detection technology are being used in the design of a HRM that should meet or exceed all these characteristics. A proof-of-principle HRM detector prototype has been constructed and results from tests of this prototype will be presented.

  4. Miniature triaxial metastable ionization detector for gas chromatographic trace analysis of extraterrestrial volatiles

    NASA Technical Reports Server (NTRS)

    Woeller, F. H.; Kojiro, D. R.; Carle, G. C.

    1984-01-01

    The present investigation is concerned with a miniature metastable ionization detector featuring an unconventional electrode configuration, whose performance characteristics parallel those of traditional design. The ionization detector is to be incorporated in a flight gas chromatograph (GC) for use in the Space Shuttle. The design of the detector is discussed, taking into account studies which verified the sensitivity of the detector. The triaxial design of the detector is compared with a flat-plate style. The obtained results show that the principal goal of developing a miniature, highly sensitive ionization detector for flight applications was achieved. Improved fabrication techniques will utilize glass-to-metal seals and brazing procedures.

  5. Influence of ambient humidity on the current delivered by air-vented ionization chambers revisited

    NASA Astrophysics Data System (ADS)

    Poirier, Aurélie; Douysset, Guilhem

    2006-10-01

    The influence of ambient humidity on the current delivered by a vented ionization chamber has been re-investigated. A Nucletron 077.091 well-type chamber together with a 192Ir HDR brachytherapy source was enclosed in a climatic test chamber and the current was recorded for various humidity values. Great care has been taken for the design of the experimental setup in order to obtain reliable measurements of currents and humidity values inside the chamber active volume. A ±0.35% linear variation of the measured currents has been observed over a common range of humidities. This result is larger than the expected variation. No formal explanation of such a discrepancy has been found yet, however the present results could lead to a set of recommendations.

  6. Correction factors for ionization chamber dosimetry in CyberKnife: Machine-specific, plan-class, and clinical fields

    SciTech Connect

    Gago-Arias, Araceli; Antolin, Elena; Fayos-Ferrer, Francisco; Simon, Rocio; Gonzalez-Castano, Diego M.; Palmans, Hugo; Sharpe, Peter; Gomez, Faustino; Pardo-Montero, Juan

    2013-01-15

    Purpose: The aim of this work is the application of the formalism for ionization chamber reference dosimetry of small and nonstandard fields [R. Alfonso, P. Andreo, R. Capote, M. S. Huq, W. Kilby, P. Kjaell, T. R. Mackie, H. Palmans, K. Rosser, J. Seuntjens, W. Ullrich, and S. Vatnitsky, 'A new formalism for reference dosimetry of small and nonstandard fields,' Med. Phys. 35, 5179-5186 (2008)] to the CyberKnife robotic radiosurgery system. Correction factors for intermediate calibration fields, a machine-specific reference field (msr) and two plan-class specific reference fields (pcsr), have been studied. Furthermore, the applicability of the new formalism to clinical dosimetry has been analyzed through the investigation of two clinical treatments. Methods: PTW31014 and Scanditronix-Wellhofer CC13 ionization chamber measurements were performed for the fields under investigation. Absorbed dose to water was determined using alanine reference dosimetry, and experimental correction factors were calculated from alanine to ionization chamber readings ratios. In addition, correction factors were calculated for the intermediate calibration fields and one of the clinical treatment fields using the Monte Carlo method and these were compared with the experimental values. Results: Overall correction factors deviating from unity by approximately 2% were obtained from both measurements and simulations, with values below and above unity for the studied intermediate calibration fields and clinical fields for the ionization chambers under consideration. Monte Carlo simulations yielded correction factors comparable with those obtained from measurements for the machine-specific reference field, although differences from 1% to 3.3% were observed between measured and calculated correction factors for the composite intermediate calibration fields. Dose distribution inhomogeneities are thought to be responsible for such discrepancies. Conclusions: The differences found between overall

  7. SU-E-T-623: Polarity Effects for Small Volume Ionization Chambers in Cobalt-60 Beams

    SciTech Connect

    Xu, Y; Bhatnagar, J; Huq, M Saiful

    2015-06-15

    Purpose: To investigate the polarity effects for small volume ionization chambers in {sup 60}Co gamma-ray beams using the Leksell Gamma Knife Perfexion. Methods: Measurements were made for 7 small volume ionization chambers (a PTW 31016, an Exradin A14, 2 Capintec PR0-5P, and 3 Exradin A16) using a PTW UNIDOSwebline Universal Dosemeter and an ELEKTA solid water phantom with proper inserts. For each ion chamber, the temperature/pressure corrected electric charge readings were obtained for 16 voltage values (±50V, ±100V, ±200V, ±300V, ±400V, ±500V, ±600V, ±700V). For each voltage, a five-minute leakage charge reading and a series of 2-minute readings were continuously taken during irradiation until 5 stable signals (less than 0.05% variation) were obtained. The average of the 5 reading was then used for the calculation of the polarity corrections at the voltage and for generating the saturation curves. Results: The polarity effects are more pronounced at high or low voltages than at the medium voltages for all chambers studied. The voltage dependence of the 3 Exradin A16 chambers is similar in shape. The polarity corrections for the Exradin A16 chambers changes rapidly from about 1 at 500V to about 0.98 at 700V. The polarity corrections for the 7 ion chambers at 300V are in the range from 0.9925 (for the PTW31016) to 1.0035 (for an Exradin A16). Conclusion: The polarity corrections for certain micro-chambers are large even at normal operating voltage.

  8. Towards reference dosimetry for the MR-linac: magnetic field correction of the ionization chamber reading

    NASA Astrophysics Data System (ADS)

    Smit, K.; van Asselen, B.; Kok, J. G. M.; Aalbers, A. H. L.; Lagendijk, J. J. W.; Raaymakers, B. W.

    2013-09-01

    In the UMC Utrecht a prototype MR-linac has been installed. The system consists of a 6 MV Elekta (Crawley, UK) linear accelerator and a 1.5 T Philips (Best, The Netherlands) Achieva MRI system. This paper investigates the feasibility to correct the ionization chamber reading for the magnetic field within the dosimetry calibration method described by Almond et al (1999 Med. Phys. 26 1847-70). Firstly, the feasibility of using an ionization chamber in an MR-linac was assessed by investigating possible influences of the magnetic field on NE2571 Farmer-type ionization chamber characteristics: linearity, repeatability, orientation in the magnetic field; and AAPM TG51 correction factor for voltage polarity and ion recombination. We found that these AAPM correction factors for the NE2571 chamber were not influenced by the magnetic field. Secondly, the influence of the permanent 1.5 T magnetic field on the NE2571 chamber reading was quantified. The reading is influenced by the magnetic field; therefore, a correction factor has been added. For the standardized setup used in this paper, the NE2571 chamber reading increases by 4.9% (± 0.2%) due to the transverse 1.5 T magnetic field. Dosimetry measurements in an MR-linac are feasible, if a setup-specific magnetic field correction factor (P1.5 T) for the charge reading is introduced. For the setup investigated in this paper, the P1.5 T has a value of 0.953.

  9. Particle and energy dependence of the statistical fluctuations of an ionization chamber current

    NASA Astrophysics Data System (ADS)

    Purghel, Lidia; Vaˆlcov, Nicolae

    For the purpose of getting more detailed information concerning the processes leading to statistical fluctuations of an ionization chamber current, measurements with various radioactive sources have been done. By using the experimental arrangement described elsewhere [A. Necula et al. Nucl. Instr. and Meth. A 332 (1993) 501] the mean value and the standard deviation of the ionization current for 3H (water vapours), 60Co (sealed source), 85Kr (gas), 204Tl (8 mm diameter disk) and 239Pu (10 mm diameter disk), beta, gamma and alpha sources have been measured. A statistical model explaining the experimental data is proposed.

  10. A new, sensitive, high resolution Raman detector based on ionization

    NASA Astrophysics Data System (ADS)

    Smith, B. W.; Omenetto, N.; Winefordner, J. D.

    1989-01-01

    A novel detection method for weak pulsed or cw Raman fluxes is described. The detector is based upon the production of Raman scatter with a tunable pulsed or cw dye laser, collecting a large fraction of the Raman scatter and transferring it efficiently into an ionization detector containing a metal (M) vapor, such as Li. The resonance ionization detector (RID) is simultaneously illuminated by a second dye laser. When the second laser is tuned to an excited state transition of the metal vapor M and when the first laser is at such a wavelength that the Raman scatter appears at the ground state absorption transition of the metal, then a current will be produced in the RID which is proportional to the Raman scatter intensity. Both the production and collection of this current can be made very efficient (approaching 100%) and should result in improved sensitivity compared to conventional dispersive or FT Raman techniques. The new approach should be much less sensitive to scatter, should have a spectral resolution better than 0.1 cm -1 and should allow Raman scatter measurements to be made at wavenumbers below 100 cm -1 and under certain conditions to 0.01 cm -1. The approach should be especially useful in highly scattering environments like Ag-sols in surface enhanced Raman and should be useful for detection of ultratrace levels of drugs and metabolites in biological fluids. The Raman-RID approach should also be useful for resonance Raman since laser scatter and molecular fluorescence should have minimal effects.

  11. Calibration of PICO Bubble Chamber Dark Matter Detectors

    NASA Astrophysics Data System (ADS)

    Jin, Miaotianzi; PICO Collaboration

    2016-03-01

    The PICO Collaboration builds bubble chambers for the direct detection of WIMP dark matter. I will present the suite of calibration experiments performed to measure the sensitivity of these chambers to nuclear recoils (the expected WIMP signal) and to gamma rays (a common background to the WIMP signal). These calibrations include measurements with a 10-ml C3F8 bubble chamber at Northwestern University and with a 30-ml C3F8 bubble chamber deployed in the University of Montreal's tandem Van de Graaf facility, giving the bubble chamber response to a variety of gamma rays, broad-spectrum neutron sources, and mono-energetic low energy neutrons. I will compare our measured sensitivities to those predicted by a simple thermodynamic model and will show how the results impact our ability to detect dark matter, with a focus on light WIMP searches. Supported by DOE Grant: DE-SC0012161.

  12. Ionization signals from diamond detectors in fast-neutron fields

    NASA Astrophysics Data System (ADS)

    Weiss, C.; Frais-Kölbl, H.; Griesmayer, E.; Kavrigin, P.

    2016-09-01

    In this paper we introduce a novel analysis technique for measurements with single-crystal chemical vapor deposition (sCVD) diamond detectors in fast-neutron fields. This method exploits the unique electronic property of sCVD diamond sensors that the signal shape of the detector current is directly proportional to the initial ionization profile. In fast-neutron fields the diamond sensor acts simultaneously as target and sensor. The interaction of neutrons with the stable isotopes 12 C and 13 C is of interest for fast-neutron diagnostics. The measured signal shapes of detector current pulses are used to identify individual types of interactions in the diamond with the goal to select neutron-induced reactions in the diamond and to suppress neutron-induced background reactions as well as γ-background. The method is verified with experimental data from a measurement in a 14.3 MeV neutron beam at JRC-IRMM, Geel/Belgium, where the 13C(n, α)10Be reaction was successfully extracted from the dominating background of recoil protons and γ-rays and the energy resolution of the 12C(n, α)9Be reaction was substantially improved. The presented analysis technique is especially relevant for diagnostics in harsh radiation environments, like fission and fusion reactors. It allows to extract the neutron spectrum from the background, and is particularly applicable to neutron flux monitoring and neutron spectroscopy.

  13. A novel convolution-based approach to address ionization chamber volume averaging effect in model-based treatment planning systems

    NASA Astrophysics Data System (ADS)

    Barraclough, Brendan; Li, Jonathan G.; Lebron, Sharon; Fan, Qiyong; Liu, Chihray; Yan, Guanghua

    2015-08-01

    The ionization chamber volume averaging effect is a well-known issue without an elegant solution. The purpose of this study is to propose a novel convolution-based approach to address the volume averaging effect in model-based treatment planning systems (TPSs). Ionization chamber-measured beam profiles can be regarded as the convolution between the detector response function and the implicit real profiles. Existing approaches address the issue by trying to remove the volume averaging effect from the measurement. In contrast, our proposed method imports the measured profiles directly into the TPS and addresses the problem by reoptimizing pertinent parameters of the TPS beam model. In the iterative beam modeling process, the TPS-calculated beam profiles are convolved with the same detector response function. Beam model parameters responsible for the penumbra are optimized to drive the convolved profiles to match the measured profiles. Since the convolved and the measured profiles are subject to identical volume averaging effect, the calculated profiles match the real profiles when the optimization converges. The method was applied to reoptimize a CC13 beam model commissioned with profiles measured with a standard ionization chamber (Scanditronix Wellhofer, Bartlett, TN). The reoptimized beam model was validated by comparing the TPS-calculated profiles with diode-measured profiles. Its performance in intensity-modulated radiation therapy (IMRT) quality assurance (QA) for ten head-and-neck patients was compared with the CC13 beam model and a clinical beam model (manually optimized, clinically proven) using standard Gamma comparisons. The beam profiles calculated with the reoptimized beam model showed excellent agreement with diode measurement at all measured geometries. Performance of the reoptimized beam model was comparable with that of the clinical beam model in IMRT QA. The average passing rates using the reoptimized beam model increased substantially from 92.1% to

  14. PTRAC File Utilization for Calculation of Free-Air Ionization Chamber Correction Factors by MCNPX

    NASA Astrophysics Data System (ADS)

    Šolc, Jaroslav; Sochor, Vladimír

    2014-06-01

    A free-air ionization chamber is used as a standard of photon air-kerma. Several correction factors are applied to the air-kerma value. Correction factors for electron loss (kloss) and for additional ionization current caused by photon scatter (ksc), photon fluorescence (kfl), photon transmission through diaphragm edge (kdtr), and photon scatter from the surface of the diaphragm aperture (kdsc) were determined by the MCNPX code utilizing information stored in Particle Track (PTRAC) output files. Individual steps of the procedure are described and the calculated values of the correction factors are presented. The values are in agreement with the correction factors published in a literature for similar free-air chambers.

  15. Some specific features of ionization chamber calibrations in linac x-ray beams at the LNE-LNHB.

    PubMed

    Delaunay, F; Ostrowsky, A

    2007-05-07

    The purpose of this note is to give some details about the modus operandi employed today to calibrate ionization chambers in radiotherapy linac photon beams at the Laboratoire National Henri Becquerel (LNE-LNHB). Some specific features are described: first the equipment (including the external monitoring ionization chambers), second the calculations of the profile or radial non-uniformity correction factors (up to 0.5% effect for commonly used ionization chambers) and finally the calculations to get the calibration coefficients for customer beam qualities.

  16. Differences between signal currents for both polarities of applied voltages on cavity ionization chambers

    NASA Astrophysics Data System (ADS)

    Takata, N.

    2001-06-01

    A difference between the surface potential of the charge collecting electrode and that of the guard electrode of an ionization chamber changes the charge collecting volume depending on the applied voltage. If the difference is large, the saturation curve of the signal current shows a maximum at a low applied voltage. Even when there is no electrical or mechanical defect, the signal current from a parallel plate ionization chamber irradiated with 60Co γ-rays increases or decreases with the applied voltage beyond the recombination region depending on the polarity of the applied voltage. The variation in the signal current is explained as a result of the change in the stopping power of air due to the acceleration or deceleration of secondary electrons. These electrons are emitted from the polarizing electrode towards the collector as a result of Compton scattering. In a range of low applied voltages, the signal current from a cylindrical ionization chamber is expected to be smaller for a negative applied voltage than for a positive applied voltage. This is because epithermal electrons are expected to have a higher probability of being lost by back diffusion than positive ions which are originally produced in a thermal equilibrium condition. An experimental result, however, showed no difference in the polarities of the applied voltage. The result may be explained as a consequence of the fact that epithemal electrons do not drift for long distances and maintain their energies.

  17. XIth Workshop on Resistive Plate Chambers and Related Detectors

    NASA Astrophysics Data System (ADS)

    RPC2012 continues the tradition of regular scientific meetings, first started in 1991 at Lecce, with following editions in Roma, Pavia, Napoli, Bari, Coimbra, Clermont-Ferrand, Seoul, Mumbai, Darmstadt. The RPC detection technique has found large application in LHC detectors and in astroparticle physics experiments. The workshop focused on performance of large RPC systems, ageing and detector materials studies, electronics for RPCs, physics of basic processes in the gas, timing applications, new ideas for RPC detectors.

  18. A numerical model of initial recombination for high-LET irradiation: Application to liquid-filled ionization chambers

    NASA Astrophysics Data System (ADS)

    Aguiar, P.; Pardo-Montero, J.

    2016-02-01

    In this paper we present a numerical model of initial recombination in media irradiated with high linear energy transfer (LET) ions, which relies on an amorphous track model of ionization of high LET particles, and diffusion, drift and recombination of ionized charge carriers. The model has fundamental applications for the study of recombination in non-polar liquids, as well as practical ones, like in modelling hadrontherapy dosimetry with ionization chambers. We have used it to study the response of liquid-filled ionization chambers to hadrontherapy beams: dependence of initial recombination on ion species, energy and applied external electric field.

  19. Ionization detector, electrode configuration and single polarity charge detection method

    DOEpatents

    He, Zhong

    1998-01-01

    An ionization detector, an electrode configuration and a single polarity charge detection method each utilize a boundary electrode which symmetrically surrounds first and second central interlaced and symmetrical electrodes. All of the electrodes are held at a voltage potential of a first polarity type. The first central electrode is held at a higher potential than the second central or boundary electrodes. By forming the first and second central electrodes in a substantially interlaced and symmetrical pattern and forming the boundary electrode symmetrically about the first and second central electrodes, signals generated by charge carriers are substantially of equal strength with respect to both of the central electrodes. The only significant difference in measured signal strength occurs when the charge carriers move to within close proximity of the first central electrode and are received at the first central electrode. The measured signals are then subtracted and compared to quantitatively measure the magnitude of the charge.

  20. Ionization detector, electrode configuration and single polarity charge detection method

    DOEpatents

    He, Z.

    1998-07-07

    An ionization detector, an electrode configuration and a single polarity charge detection method each utilize a boundary electrode which symmetrically surrounds first and second central interlaced and symmetrical electrodes. All of the electrodes are held at a voltage potential of a first polarity type. The first central electrode is held at a higher potential than the second central or boundary electrodes. By forming the first and second central electrodes in a substantially interlaced and symmetrical pattern and forming the boundary electrode symmetrically about the first and second central electrodes, signals generated by charge carriers are substantially of equal strength with respect to both of the central electrodes. The only significant difference in measured signal strength occurs when the charge carriers move to within close proximity of the first central electrode and are received at the first central electrode. The measured signals are then subtracted and compared to quantitatively measure the magnitude of the charge. 10 figs.

  1. Application of the two-dose-rate method for general recombination correction for liquid ionization chambers in continuous beams

    NASA Astrophysics Data System (ADS)

    Andersson, Jonas; Tölli, Heikki

    2011-01-01

    A method to correct for the general recombination losses for liquid ionization chambers in continuous beams has been developed. The proposed method has been derived from Greening's theory for continuous beams and is based on measuring the signal from a liquid ionization chamber and an air filled monitor ionization chamber at two different dose rates. The method has been tested with two plane parallel liquid ionization chambers in a continuous radiation x-ray beam with a tube voltage of 120 kV and with dose rates between 2 and 13 Gy min-1. The liquids used as sensitive media in the chambers were isooctane (C8H18) and tetramethylsilane (Si(CH3)4). The general recombination effect was studied using chamber polarizing voltages of 100, 300, 500, 700 and 900 V for both liquids. The relative standard deviation of the results for the collection efficiency with respect to general recombination was found to be a maximum of 0.7% for isooctane and 2.4% for tetramethylsilane. The results are in excellent agreement with Greening's theory for collection efficiencies over 90%. The measured and corrected signals from the liquid ionization chambers used in this work are in very good agreement with the air filled monitor chamber with respect to signal to dose linearity.

  2. Application of the two-dose-rate method for general recombination correction for liquid ionization chambers in continuous beams.

    PubMed

    Andersson, Jonas; Tölli, Heikki

    2011-01-21

    A method to correct for the general recombination losses for liquid ionization chambers in continuous beams has been developed. The proposed method has been derived from Greening's theory for continuous beams and is based on measuring the signal from a liquid ionization chamber and an air filled monitor ionization chamber at two different dose rates. The method has been tested with two plane parallel liquid ionization chambers in a continuous radiation x-ray beam with a tube voltage of 120 kV and with dose rates between 2 and 13 Gy min(-1). The liquids used as sensitive media in the chambers were isooctane (C(8)H(18)) and tetramethylsilane (Si(CH(3))(4)). The general recombination effect was studied using chamber polarizing voltages of 100, 300, 500, 700 and 900 V for both liquids. The relative standard deviation of the results for the collection efficiency with respect to general recombination was found to be a maximum of 0.7% for isooctane and 2.4% for tetramethylsilane. The results are in excellent agreement with Greening's theory for collection efficiencies over 90%. The measured and corrected signals from the liquid ionization chambers used in this work are in very good agreement with the air filled monitor chamber with respect to signal to dose linearity.

  3. Calibration of the KRISS reference ionization chamber for certification of ²²²Rn gaseous sources.

    PubMed

    Lee, J M; Lee, K B; Lee, S H; Oh, P J; Park, T S; Kim, B C; Lee, M S

    2013-11-01

    A primary measurement system for gaseous (222)Rn based on the defined solid angle counting method has recently been constructed at KRISS and the reference ionization chamber used to measure the activities of gamma-emitting single radionuclides was adopted as a secondary standard for gaseous (222)Rn. A 20 mL flame-sealed glass ampoule source from the primary measurement system was used to calibrate the ionization chamber for (222)Rn. The (222)Rn efficiency of the ionization chamber was compared with that calculated by using a photon energy-dependent efficiency curve and that measured by using a standard (226)Ra solution. From the comparisons we draw the conclusion that the reference ionization chamber for gamma-emitting radionuclides can be a suitable secondary measurement system for gaseous (222)Rn sources.

  4. WE-EF-207-11: Energy and Depth Response of Thermoluminescent Dosimeters and Ionization Chambers in Water for Kilovoltage X-Ray Beams

    SciTech Connect

    Lawless, M; Palmer, B; DeWerd, L

    2015-06-15

    Purpose: To assess the effects of changes in beam quality on detector response in the kilovoltage energy range by modulating the x-ray tube voltage and the measurement depth in water. Methods: Measurements were performed with TLD-100 and TLD-100H thermoluminescent dosimeters and an A12 farmer-type ionization chamber. To assess the energy response of the detectors, irradiations were performed at a depth of 3 cm in a custom-built thin-window water phantom using the moderately filtered x-ray beams at the UWADCL (20 kVp-250 kVp) and a Co-60 beam.The x-ray beams and detectors were modeled using the EGSnrc Monte Carlo code. The model was validated by simulating dose to the collecting volume of an A12 farmer chamber and comparing it with measured A12 signal as a function of depth. Dose was tallied to each detector and to water for comparison with measurements. Simulations were used to calculate the predicted energy response, which was compared to the measured response of each detector. Dose to each detector and dose to water as a function of depth were also simulated. Results: Detector output per dose to water was found to deviate by up to 15%, 20% and 30% as a function of energy relative to Co-60 for the A12, TLD-100H and TLD-100, respectively. The EGSnrc simulations produced results similar to the measurements for ionization chambers, but discrepancies of up to 30% were observed for TLD-100H. Simulated detector response as a function of depth was found to vary by up to 3%. Conclusion: These results suggest that changes in beam quality in kilovoltage x-ray beams can have a significant impact on detector response. In-water detector response was found to differ from the previously investigated in-air response. Deviations in detector response as a function of depth were less significant, but could potentially cause dosimetric errors if ignored.

  5. A two-dimensional liquid-filled ionization chamber array prototype for small-field verification: characterization and first clinical tests

    NASA Astrophysics Data System (ADS)

    Brualla-González, Luis; Gómez, Faustino; Vicedo, Aurora; González-Castaño, Diego M.; Gago-Arias, Araceli; Pazos, Antonio; Zapata, Martín; Roselló, Joan V.; Pardo-Montero, Juan

    2012-08-01

    In this work we present the design, characterization and first clinical tests of an in-house developed two-dimensional liquid-filled ionization chamber prototype for the verification of small radiotherapy fields and treatments containing such small fields as in radiosurgery, which consists of 2 mm × 2 mm pixels arranged on a 16×8 rectangular grid. The ionization medium is isooctane. The characterization of the device included the study of depth, field-size and dose-rate dependences, which are sufficiently moderate for a good operation at therapy radiation levels. However, the detector presents an important anisotropic response, up to ≃ 12% for front versus near-lateral incidence, which can impact the verification of full treatments with different incidences. In such a case, an anisotropy correction factor can be applied. Output factors of small square fields measured with the device show a small systematic over-response, less than 1%, when compared to unshielded diode measurements. An IMRT radiosurgery treatment has been acquired with the liquid-filled ionization chamber device and compared with film dosimetry by using the gamma method, showing good agreement: over 99% passing rates for 1.2% and 1.2 mm for an incidence-per-incidence analysis; 100% passing rates for tolerances 1.8% and 1.8 mm when the whole treatment is analysed and the anisotropy correction factor is applied. The point dose verification for each incidence of the treatment performed with the liquid-filled ionization chamber agrees within 1% with a CC01 ionization chamber. This prototype has shown the utility of this kind of technology for the verification of small fields/treatments. Currently, a larger device covering a 5 cm × 5 cm area is under development.

  6. The calibration and use of plane-parallel ionization chambers for dosimetry of electron beams.

    PubMed

    Almond, P R; Xu, Z; Li, H; Park, H C

    1995-08-01

    The AAPM TG 39 protocol has proposed three different methods of calibrating plane-parallel ionization chambers, i.e., in-phantom irradiation with a high-energy electron beam and in-phantom and in-air 60Co irradiation. To verify the consistency of the three methods, we have measured Ngaspp values using each of these techniques for the five most commonly used plane-parallel chambers considered by the protocol. Our results demonstrate that the measured Ngaspp values for the three different methods for any of the chambers agree to within +/- 0.6%. Once Ngaspp was measured, the determination of absorbed dose for electron beams with different energies for an AECL Therac 20 and Philips SL25 was carried out according to the AAPM TG 39 protocol. The results show that the determination of the absorbed dose outputs for any of the five chambers agree to within +/- 0.7% for electron-beam energies of 4-20 MeV if all five chambers had Ngaspp values determined by the electron-beam method. The uncertainties are well within the expected error for these approaches.

  7. The response of prototype plane-parallel ionization chambers in small megavoltage x-ray fields.

    PubMed

    McNiven, Andrea L; Mulligan, Matt; Kron, Tomas; Battista, Jerry J

    2006-11-01

    Accurate small-field dosimetry has become important with the use of multiple small fields in modern radiotherapy treatments such as IMRT and stereotactic radiosurgery. In this study, we investigate the response of a set of prototype plane-parallel ionization chambers, based upon the Exradin T11 chamber, with active volume diameters of 2, 4, 10, and 20 mm, exposed to 6 MV stereotactic radiotherapy x-ray fields. Our goal was to assess their usefulness for accurate small x-ray field dose measurements. The relative ionization response was measured in circular fields (0.5 to 4 cm diameter) as compared to a 10 x 10 cm2 reference field. A large discrepancy (approximately 40%) was found between the relative response in the smallest plane-parallel chamber and other small volume dosimeters (radiochromic film, micro-metal-oxide-semiconductor field-effect transistor and diode) used for comparison. Monte Carlo BEAMnrc simulations were used to simulate the experimental setup in order to investigate the cause of the under-response and to calculate appropriate correction factors that could be applied to experimental measurements. It was found that in small fields, the air cavity of these custom-made research chambers perturbed the secondary electron fluence profile significantly, resulting in decreased fluence within the active volume, which in turn produces a chamber under-response. It is demonstrated that a large correction to the p(fl) correction factor would be required to improve dosimetric accuracy in small fields, and that these factors could be derived using Monte Carlo simulations.

  8. Characterization and performances of a monitoring ionization chamber dedicated to IBA-universal irradiation head for Pencil Beam Scanning

    NASA Astrophysics Data System (ADS)

    Courtois, C.; Boissonnat, G.; Brusasco, C.; Colin, J.; Cussol, D.; Fontbonne, J. M.; Marchand, B.; Mertens, T.; de Neuter, S.; Peronnel, J.

    2014-02-01

    Every radiotherapy center has to be equipped with real-time beam monitoring devices. In 2008, we developed an ionization chamber in collaboration with the IBA (Ion Beam Applications) company. This monitoring device called IC2/3 was developed to be used in IBA universal irradiation head for Pencil Beam Scanning (PBS). Here we present the characterization of the IC2/3 monitor in the energy and flux ranges used in protontherapy. The equipment has been tested with an IBA cyclotron able to deliver proton beams from 70 to 230 MeV. This beam monitoring device has been validated and is now installed at the Westdeutsches Protonentherapiezentrum Essen protontherapy center (WPE, Germany). The results obtained in both terms of spatial resolution and dose measurements are at least equal to the initial specifications needed for PBS purposes. The detector measures the dose with a relative uncertainty lower than 1% in the range from 0.5 Gy/min to 8 Gy/min while the spatial resolution is better than 250 μm. The technology has been patented and five IC2/3 chambers were delivered to IBA. Nowadays, IBA produces the IC2/3 beam monitoring device as a part of its Proteus 235 product.

  9. Smoke Emissions from Aircraft Interior Materials at Elevated Heat Flux Levels Using Modified NBS Smoke Chamber.

    DTIC Science & Technology

    1979-07-01

    time at each test condition was calculated and plotted at 0.1-minute interalonaHwetPcrd2O computer-plotter (appendix B). For 2.2 Btu/ft’ a piloted and...nospiloted tests and 5.0 Btu/ft 2 s piloted tests, three samples of every material were tested. In calculating specific optical density plots, the...grouping the materials into five usage categories. FABRICS. A smoke limit once considered for fabrics was Ds C 100 at 4 minutes for a 2.2 Btu/ft 2 s

  10. Measurement of photon flux with a miniature gas ionization chamber in a Material Testing Reactor

    NASA Astrophysics Data System (ADS)

    Fourmentel, D.; Filliatre, P.; Villard, J. F.; Lyoussi, A.; Reynard-Carette, C.; Carcreff, H.

    2013-10-01

    Nuclear heating measurements in Material Testing Reactors (MTR) are crucial for the design of the experimental devices and the prediction of the temperature of the hosted samples. Nuclear heating in MTR materials (except fuel) is mainly due to the energy deposition by the photon flux. Therefore, the photon flux is a key input parameter for the computer codes which simulate nuclear heating and temperature reached by samples/devices under irradiation. In the Jules Horowitz MTR under construction at the CEA Cadarache, the maximal expected nuclear heating levels will be about 15 to 18 W g-1 and it will be necessary to assess this parameter with the best accuracy. An experiment was performed at the OSIRIS reactor to combine neutron flux, photon flux and nuclear heating measurements to improve the knowledge of the nuclear heating in MTR. There are few appropriate sensors for selective measurement of the photon flux in MTR even if studies and developments are ongoing. An experiment, called CARMEN-1, was conducted at the OSIRIS MTR and we used in particular a gas ionization chamber based on miniature fission chamber design to measure the photon flux. In this paper, we detail Monte-Carlo simulations to analyze the photon fluxes with ionization chamber measurements and we compare the photon flux calculations to the nuclear heating measurements. These results show a good accordance between photon flux measurements and nuclear heating measurement and allow improving the knowledge of these parameters.

  11. Ambiguities in the grid-inefficiency correction for Frisch-Grid Ionization Chambers

    NASA Astrophysics Data System (ADS)

    Al-Adili, A.; Hambsch, F.-J.; Bencardino, R.; Oberstedt, S.; Pomp, S.

    2012-05-01

    Ionization chambers with Frisch grids have been very successfully applied to neutron-induced fission-fragment studies during the past 20 years. They are radiation resistant and can be easily adapted to the experimental conditions. The use of Frisch grids has the advantage to remove the angular dependency from the charge induced on the anode plate. However, due to the Grid Inefficiency (GI) in shielding the charges, the anode signal remains slightly angular dependent. The correction for the GI is, however, essential to determine the correct energy of the ionizing particles. GI corrections can amount to a few percent of the anode signal. Presently, two contradicting correction methods are considered in literature. The first method adding the angular-dependent part of the signal to the signal pulse height; the second method subtracting the former from the latter. Both additive and subtractive approaches were investigated in an experiment where a Twin Frisch-Grid Ionization Chamber (TFGIC) was employed to detect the spontaneous fission fragments (FF) emitted by a 252Cf source. Two parallel-wire grids with different wire spacing (1 and 2 mm, respectively), were used individually, in the same chamber side. All the other experimental conditions were unchanged. The 2 mm grid featured more than double the GI of the 1 mm grid. The induced charge on the anode in both measurements was compared, before and after GI correction. Before GI correction, the 2 mm grid resulted in a lower pulse-height distribution than the 1 mm grid. After applying both GI corrections to both measurements only the additive approach led to consistent grid independent pulse-height distributions. The application of the subtractive correction on the contrary led to inconsistent, grid-dependent results. It is also shown that the impact of either of the correction methods is small on the FF mass distributions of 235U(nth, f).

  12. Monte Carlo calculations of correction factors for plane-parallel ionization chambers in clinical electron dosimetry

    SciTech Connect

    Araki, Fujio

    2008-09-15

    Recent standard dosimetry protocols recommend that plane-parallel ionization chambers be used in the measurements of depth-dose distributions or the calibration of low-energy electron beams with beam quality R{sub 50}<4 g/cm{sup 2}. In electron dosimetry protocols with the plane-parallel chambers, the wall correction factor, P{sub wall}, in water is assumed to be unity and the replacement correction factor, P{sub repl}, is taken to be unity for well-guarded plane-parallel chambers, at all measurement depths. This study calculated P{sub wall} and P{sub repl} for NACP-02, Markus, and Roos plane-parallel chambers in clinical electron dosimetry using the EGSnrc Monte Carlo code system. The P{sub wall} values for the plane-parallel chambers increased rapidly as a function of depth in water, especially at lower energy. The value around R{sub 50} for NACP-02 was about 10% greater than unity at 4 MeV. The effect was smaller for higher electron energies. Similarly, P{sub repl} values with depth increased drastically at the region with the steep dose gradient for lower energy. For Markus P{sub repl} departed more than 10% from unity close to R{sub 50} due to the narrow guard ring width. P{sub repl} for NACP-02 and Roos was close to unity in the plateau region of depth-dose curves that includes a reference depth, d{sub ref}. It was also found that the ratio of the dose to water and the dose to the sensitive volume in the air cavity for the plane-parallel chambers, D{sub w}/[D{sub air}]{sub pp}, at d{sub ref} differs significantly from that assumed by electron dosimetry protocols.

  13. Argon/propane ionization-chamber dosimetry for mixed x-ray/neutron fields.

    PubMed

    Schulz, R J

    1978-01-01

    The photoneutrons produced by high-energy x-ray machines can diffuse through the mazes usually employed at the treatment-room entrance and readily penetrate the lead-lined doors used for x-ray shielding. The measurement of these neutrons in the presence of x-rays and the determination of dose equivalent poses a problem for which there is currently no standard method of solution. In order to separate x-ray dose from neutron dose, the author employed an ionization chamber alternately filled with argon or propane. The response characteristics of this chamber to x-ray and neutrons are described. Quality factors were determined from a calculated neutron spectrum. As a result of these measurements, a 10-in. polyethylene door was added to the entranceway of a 25-MV linear accelerator.

  14. Feasibility of calibrating elongated brachytherapy sources using a well-type ionization chamber

    SciTech Connect

    Meigooni, Ali S.; Awan, Shahid B.; Dou, Kai

    2006-11-15

    Recently, elongated brachytherapy sources (active length >1 cm) have become commercially available for interstitial prostate implants. These sources were introduced to improve the quality of brachytherapy procedures by eliminating the migration and seed bunching associated with loose seed-type implants. However, the inability to calibrate elongated brachytherapy sources with the Wide-Angle Free-Air Chamber (WAFAC) used by the National Institute of Standards and Technology (NIST) hinders the experimental determination of dosimetric parameters of these source types. In order to resolve this shortcoming, an interim solution has been introduced for calibration of elongated brachytherapy sources using a commercially available well-type ionization chamber. The feasibility of this procedure was examined by calibrating RadioCoil{sup Tm} {sup 103}Pd sources with active lengths ranging from 1 to 7 cm.

  15. Pulsed Discharge Helium Ionization Detector for Highly Sensitive Aquametry.

    PubMed

    Mowry, Curtis D; Pimentel, Adam S; Sparks, Elizabeth S; Moorman, Matthew W; Achyuthan, Komandoor E; Manginell, Ronald P

    2016-01-01

    Trace moisture quantitation is crucial in medical, civilian and military applications. Current aquametry technologies are limited by the sample volume, reactivity, or interferences, and/or instrument size, weight, power, cost, and complexity. We report for the first time on the use of a pulsed discharge helium ionization detector (PDHID-D2) (∼196 cm(3)) for the sensitive (limit of detection, 0.047 ng; 26 ppm), linear (r(2) >0.99), and rapid (< 2 min) quantitation of water using a small (0.2 - 5.0 μL) volume of liquid or gas. The relative humidity sensitivity was 0.22% (61.4 ppmv) with a limit of detection of less than 1 ng moisture with gaseous samples. The sensitivity was 10 to 100 to fold superior to competing technologies without the disadvantages inherent to these technologies. The PDHID-D2, due to its small footprint and low power requirement, has good size, weight, and power-portability (SWAPP) factors. The relatively low cost (∼$5000) and commercial availability of the PDHID-D2 makes our technique applicable to highly sensitive aquametry.

  16. Technical Note: Experimental determination of the effective point of measurement of two cylindrical ionization chambers in a clinical proton beam

    SciTech Connect

    Sugama, Yuya; Nishio, Teiji; Onishi, Hiroshi

    2015-07-15

    Purpose: IAEA TRS-398 notes that cylindrical ionization chambers are preferred for reference proton dosimetry. If a cylindrical ionization chamber is used in a phantom to measure the dose as a function of depth, the effective point of measurement (EPOM) must be taken into account. IAEA TRS-398 recommends a displacement of 0.75 times the inner cavity radius (0.75R) for heavy ion beams. Theoretical models by Palmans and by Bhullar and Watchman confirmed this value. However, the experimental results vary from author to author. The purpose of this study is to accurately measure the displacement and explain the past experimental discrepancies. Methods: In this work, we measured the EPOM of cylindrical ionization chambers with high accuracy by comparing the Bragg-peak position obtained with cylindrical ionization chambers (PTW 30013, PTW 31016) to that obtained using a plane-parallel ionization chamber (PTW 34045). Results: The EPOMs of PTW 30013 and 31016 were shifted by 0.92 ± 0.07 R with R = 3.05 mm and 0.90 ± 0.14 R with R = 1.45 mm, respectively, from the reference point toward the source. Conclusions: The EPOMs obtained were greater than the value of 0.75R proposed by the IAEA TRS-398 and the analytical results.

  17. Assessment of ionization chamber correction factors in photon beams using a time saving strategy with PENELOPE code.

    PubMed

    Reis, C Q M; Nicolucci, P

    2016-02-01

    The purpose of this study was to investigate Monte Carlo-based perturbation and beam quality correction factors for ionization chambers in photon beams using a saving time strategy with PENELOPE code. Simulations for calculating absorbed doses to water using full spectra of photon beams impinging the whole water phantom and those using a phase-space file previously stored around the point of interest were performed and compared. The widely used NE2571 ionization chamber was modeled with PENELOPE using data from the literature in order to calculate absorbed doses to the air cavity of the chamber. Absorbed doses to water at reference depth were also calculated for providing the perturbation and beam quality correction factors for that chamber in high energy photon beams. Results obtained in this study show that simulations with phase-space files appropriately stored can be up to ten times shorter than using a full spectrum of photon beams in the input-file. Values of kQ and its components for the NE2571 ionization chamber showed good agreement with published values in the literature and are provided with typical statistical uncertainties of 0.2%. Comparisons to kQ values published in current dosimetry protocols such as the AAPM TG-51 and IAEA TRS-398 showed maximum percentage differences of 0.1% and 0.6% respectively. The proposed strategy presented a significant efficiency gain and can be applied for a variety of ionization chambers and clinical photon beams.

  18. Matrix:. AN Innovative Pixel Ionization Chamber for On-Line Beam Monitoring in Hadrontherapy

    NASA Astrophysics Data System (ADS)

    Braccini, S.; Pitta', G.; Donetti, M.; Cirio, R.; La Rosa, A.; Garella, M. A.; Giordanengo, S.; Marchetto, F.; Peroni, C.

    2006-04-01

    The control of intensity, position and shape of clinical beams are key issues in the treatment of tumours using hadron beams, especially in the case of active dose distribution systems. For this purpose an innovative pixel ionization chamber, named MATRIX, has been designed, constructed and tested. The chamber is conceived to be located very near the patient to precisely monitor the beam parameters used to verify the treatment planning specifications. MATRIX operates in air and is characterized by a 21 × 21 cm2 sensitive area subdivided in 1024 pixels of 6.5 × 6.5 mm2. To minimize the amount of material crossed by the beam, the anode is made of a 50 μm kapton foil, with a deposit of 17 μm copper on each side. A very sensitive electronics is used for the readout, based on a dedicated chip. In this paper the construction of the chamber and the very positive results of the first beam tests are described.

  19. [Experimental investigation of the collection efficiency of a PTW Roos ionization chamber irradiated with pulsed beams at high pulse dose with different pulse lengths].

    PubMed

    Karsch, Leonhard; Richter, Christian; Pawelke, Jörg

    2011-01-01

    In gas-filled ionization chambers as radiation detectors, the collection of the charge carriers is affected by the recombination effect. In dosimetry this effect must be accounted for by the saturation correction factor k(S). The physical description of the correction factor by Boag, Hochhäuser and Balk for pulsed radiation is well established. However, this description is only accurate when the pulse length is short compared to the collection time of the ionization chamber. In this work experimental investigations of the saturation correction factor have been made for pulses of 4 μ s up to pulse doses of about 230 mGy, and the theory of Boag, Hochhäuser and Balk was again confirmed. For longer pulses, however, the correction factor decreases and at a pulse duration of about 200μs reaches 75% of the value valid for short pulses. This reduced influence of the ion recombination is interpreted by the reaction kinetics of ion recombination as a second-order reaction. This effect is negligible for PTW Roos chambers at clinical linear accelerators with 4 μ s pulse duration for pulse doses up to 120 mGy.

  20. Determination of small-field correction factors for cylindrical ionization chambers using a semiempirical method

    NASA Astrophysics Data System (ADS)

    Park, Kwangwoo; Bak, Jino; Park, Sungho; Choi, Wonhoon; Park, Suk Won

    2016-02-01

    A semiempirical method based on the averaging effect of the sensitive volumes of different air-filled ionization chambers (ICs) was employed to approximate the correction factors for beam quality produced from the difference in the sizes of the reference field and small fields. We measured the output factors using several cylindrical ICs and calculated the correction factors using a mathematical method similar to deconvolution; in the method, we modeled the variable and inhomogeneous energy fluence function within the chamber cavity. The parameters of the modeled function and the correction factors were determined by solving a developed system of equations as well as on the basis of the measurement data and the geometry of the chambers. Further, Monte Carlo (MC) computations were performed using the Monaco® treatment planning system to validate the proposed method. The determined correction factors (k{{Q\\text{msr}},Q}{{f\\text{smf}}, {{f}\\text{ref}}} ) were comparable to the values derived from the MC computations performed using Monaco®. For example, for a 6 MV photon beam and a field size of 1  ×  1 cm2, k{{Q\\text{msr}},Q}{{f\\text{smf}}, {{f}\\text{ref}}} was calculated to be 1.125 for a PTW 31010 chamber and 1.022 for a PTW 31016 chamber. On the other hand, the k{{Q\\text{msr}},Q}{{f\\text{smf}}, {{f}\\text{ref}}} values determined from the MC computations were 1.121 and 1.031, respectively; the difference between the proposed method and the MC computation is less than 2%. In addition, we determined the k{{Q\\text{msr}},Q}{{f\\text{smf}}, {{f}\\text{ref}}} values for PTW 30013, PTW 31010, PTW 31016, IBA FC23-C, and IBA CC13 chambers as well. We devised a method for determining k{{Q\\text{msr}},Q}{{f\\text{smf}}, {{f}\\text{ref}}} from both the measurement of the output factors and model-based mathematical computation. The proposed method can be useful in case the MC simulation would not be applicable for the clinical settings.

  1. Determination of polycyclic aromatic hydrocarbons [PAHs] in processed meat products using gas chromatography - flame ionization detector.

    PubMed

    Olatunji, Olatunde S; Fatoki, Olalekan S; Opeolu, Beatrice O; Ximba, Bhekumusa J

    2014-08-01

    The concentrations of polycyclic aromatic hydrocarbons (PAHs) in smoked, grilled and boiled meats were determined using gas chromatography - flame ionization detector (GC-FID). PAHs in the processed meats were extracted in n-hexane after hydrolysis with methanolic KOH. Clean-up was achieved using solid phase extraction in neutral-Si/basic-Si/acidic-Si/neutral-Si frits. The fractions, benzo[k]fluoranthene (BkP), benzo[a]pyrene (BaP), indeno[123-cd]pyrene (IP) and benzo[ghi]perylene (BghiP) were separated and quantified using GC-FID. The method and instrument limits of detections were 0.1, 0.1, 0.2, 0.3μg/kg and 0.5, 0.5, 1.0, 1.5μg/kg, respectively, for BkP, BaP, IP and BghiP. The method's recovery and precision generally varied between 83.69% and 94.25% with relative standard deviation (RSD) of 3.18-15.60%; and 90.38-96.71% with relative standard deviation (RSD) of 1.82-12.87% respectively. The concentration of BkP, BaP, IP and BghiP in smoked, grilled and boiled meat samples were ranged 0.64-31.54μg/kg, 0.07-7.04μg/kg, 0.09-15.03, 0.51-46.67μg/kg and 0.01-5.11μg/kg, respectively.

  2. Performance of a multi-axis ionization chamber array in a 1.5 T magnetic field

    NASA Astrophysics Data System (ADS)

    Smit, K.; Kok, J. G. M.; Lagendijk, J. J. W.; Raaymakers, B. W.

    2014-04-01

    At the UMC Utrecht a prototype MR-linac has been installed. The system consists of an 8 MV Elekta linear accelerator and a 1.5 T Philips MRI system. This paper investigates the performance of the IC PROFILER™, a multi-axis ionization chamber array, in a 1.5 T magnetic field. The influence of the magnetic field on the IC PROFILER™ reproducibility, dose response linearity, pulse rate frequency dependence, power to electronics, panel orientation and ionization chamber shape were investigated. The linearity, reproducibility, pulse rate frequency dependence, panel orientation and ionization chamber shape are unaffected by the magnetic field. When the measurements results are normalized to the centre reference chamber, the measurements can commence unaltered. Orientation of the ionization chambers in the magnetic field is of importance, therefore caution must be taken when comparing or normalizing results from several different axes. IC PROFILER™ dose profiles were compared with film dose profiles obtained simultaneously in the MR-linac. Deviation between the film and the IC PROFILER™ data was caused by the noise in the film, indicating correct performance of the IC PROFILER™ in the transverse 1.5 T magnetic field.

  3. [Theoretical investigation of the saturation correction for ionization chambers irradiated with pulsed beams of arbitrary pulse length].

    PubMed

    Karsch, Leonhard; Pawelke, Jörg

    2014-09-01

    In ionization chambers, not all released charge is collected due to the recombination of charge carriers. This effect is taken into account by the saturation correction factor kS. A physical description of the correction factor has been established for pulsed radiation. However, it is only accurate when the pulse length is short compared with the collection time of the ionization chamber. In this paper we develop a description of the saturation correction for radiation pulses of arbitrary length. For this, a system of partial differential equations is solved iteratively. The numerical solutions are verified experimentally for a Roos ionization chamber (PTW TM34001) exposed to a pulsed electron beam. The results of this iterative procedure describe the experimental data well. The calculations are also possible for beam structures which are experimentally hard to get and thereby contribute to a better understanding and correct description of the saturation correction at arbitrary pulse length. Among other things the pulse length dependent distributions of the charge carriers in the ionization chamber is calculated, inclusive of the transition to the conditions prevailing in the case of continuous irradiation. Furthermore is shown that the formula for kS established by Hochhäuser and Balk is applicable even at arbitrary pulse length, if pulse duration dependent effective values are used for the parameters a and p. These effective values have been determined for the Roos chamber at pulse lengths up to 300 μs.

  4. Long-term stability of liquid ionization chambers with regard to their qualification as local reference dosimeters for low dose-rate absorbed dose measurements in water.

    PubMed

    Bahar-Gogani, J; Grindborg, J E; Johansson, B E; Wickman, G

    2001-03-01

    The long-term sensitivity and calibration stability of liquid ionization chambers (LICs) has been studied at a local and a secondary standards dosimetry laboratory over a period of 3 years. The chambers were transported several times by mail between the two laboratories for measurements. The LICs used in this work are designed for absorbed dose measurements in the dose rate region of 0.1-100 mGy min(-1) and have a liquid layer thickness of 1 mm and a sensitive volume of 16.2 mm3. The liquids used as sensitive media in the chambers are mixtures of isooctane (C8H18) and tetramethylsilane (Si(CH3)4) in different proportions (about 2 to 1). Operating at a polarizing voltage of 300 V the leakage current of the chambers was stable and never exceeded 3% of the observable current at a dose rate of about 1 mGy min(-1). The volume sensitivity of the chambers was measured to be of the order of 10(-9) C Gy(-1) mm3. No systematic changes in the absorbed dose to water calibration was observed for any of the chambers during the test period (sigma < 0.2%). Variations in chamber dose response with small changes in the polarizing voltage as well as sensitivity changes with accumulated absorbed dose were also investigated. Measurements showed that the LIC response varies by 0.15% per 1% change in applied voltage around 300 V. No significant change could be observed in the LIC sensitivity after a single absorbed dose of 15 kGy. The results indicate that the LIC can be made to serve as a calibration transfer instrument and a reference detector for absorbed dose to water determinations providing good precision and long-term reproducibility.

  5. Poster — Thur Eve — 21: Off-axis dose perturbation effects in water in a 5 × 5 cm{sup 2} 18 MV photon beam for the PTW microLion and Exradin A1SL ionization chambers

    SciTech Connect

    O'Grady, K; Davis, S D; Papaconstadopoulos, P; Seuntjens, J

    2014-08-15

    A PTW microLion liquid ionization chamber and an Exradin A1SL air-filled ionization chamber have been modeled using the egs-chamber user code of the EGSnrc system to determine their perturbation effects in water in a 5 × 5 cm{sup 2} 18 MV photon beam. A model of the Varian CL21EX linear accelerator was constructed using the BEAMnrc Monte Carlo code, and was validated by comparing measured PDDs and profiles from the microLion and A1SL chambers to calculated results that included chamber models. Measured PDDs for a 5 × 5 cm{sup 2} field for the microLion chamber agreed with calculations to within 1.5% beyond a depth of 0.5 cm, and the A1SL PDDs agreed within 1.0% beyond 1.0 cm. Measured and calculated profiles at 10 cm depth agreed within 1.0% for both chambers inside the field, and within 4.0% near the field edge. Local percent differences increased up to 15% at 4 cm outside the field. The ratio of dose to water in the absence of the chamber relative to dose in the chamber's active volume as a function of off-axis distance was calculated using the egs-chamber correlated sampling technique. The dose ratio was nearly constant inside the field and consistent with the stopping power ratios of water to detector material, but varied up to 3.3% near the field edge and 5.2% at 4 cm outside the field. Once these perturbation effects are fully characterized for more field sizes and detectors, they could be applied to clinical water tank measurements for improved dosimetric accuracy.

  6. Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories.

    PubMed

    Hourdakis, Costas J; Büermann, Ludwig; Ciraj-Bjelac, Olivera; Csete, Istvan; Delis, Harry; Gomola, Igor; Persson, Linda; Novak, Leos; Petkov, Ivailo; Toroi, Paula

    2016-01-01

    A comparison of calibration results and procedures in terms of air kerma length product, PKL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the "partial irradiation method" is the preferred method for the pencil-type IC calibration in terms of PKL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement (>99%) for both dosimetry quantities (PKL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed.

  7. Performance of a gas flow ionization detector filled with He-iso-C4H10 mixtures for STIM-T

    NASA Astrophysics Data System (ADS)

    Marques, A. C.; Fraga, M. M. F. R.; Fonte, P.; Beasley, D. G.; Cruz, C.; Alves, L. C.; da Silva, R. C.

    2015-04-01

    A cylindrical gas flow ionization chamber has been developed for measuring particle energy in Scanning Transmission Ion Microscopy Tomography (STIM-T) experiments due to its ability to withstand the direct beam. The response of a He-iso-C4H10 filled ionization detector to 2 MeV H+ and He+ beams was studied. Different operating parameters, such as concentration of isobutane (in the range of 55-100%), anode voltage, amplifier shaping time, the geometry of the detector entrance canal and the solid angle of the detector, were investigated. The stable operating plateau and the anode voltage at which the best energy resolution is attained were also determined for every gas mixture. The best energy resolution achieved so far for 2 MeV H+ and He+ static beams was ∼1.3%, which is comparable to that of Si PIN diode detectors (in the range of 15-30 keV). Computed tomography (CT) was applied to a set of STIM projections acquired with the gas ionization chamber at the IST/CTN microprobe beam line in order to visualize the 3D-mass distribution in a test structure.

  8. Unexpected bias in NIST 4πγ ionization chamber measurements.

    PubMed

    Unterweger, M P; Fitzgerald, R

    2012-09-01

    In January of 2010, it was discovered that the source holder used for calibrations in the NIST 4πγ ionization chamber (IC) has not been stable. The positioning ring that determines the height of the sample in the reentrant tube of the IC has slowly shifted during 35 years of use. This has led to a slow change in the calibration factors for the various radionuclides measured by this instrument. The changes are dependent on γ-ray energy and the time the IC was calibrated for a given radionuclide. A review of the historic data with regard to when the calibrations were done has enabled us to approximate the magnitude of the changes with time. This requires a number of assumptions, and corresponding uncertainty components, including whether the changes in height were gradual or in steps as will be shown in drawings of sample holder. For calibrations the changes in calibration factors have been most significant for low energy gamma emitters such as (133)Xe, (241)Am, (125)I and (85)Kr. The corrections to previous calibrations can be approximated and the results corrected with an increase in the overall uncertainty. At present we are recalibrating the IC based on new primary measurements of the radionuclides measured on the IC. Likewise we have been calibrating a new automated ionization-chamber system. A bigger problem is the significant number of half-life results NIST has published over the last 35 years that are based on IC measurements. The effect on half-life is largest for long-lived radionuclei, especially low-energy γ-ray emitters. This presentation will review our results and recommend changes in values and/or uncertainties. Any recommendation for withdrawal of any results will also be undertaken.

  9. Dependence with air density of the response of the PTW SourceCheck ionization chamber for low energy brachytherapy sources

    SciTech Connect

    Tornero-López, Ana M.; Guirado, Damián; Ruiz-Arrebola, Samuel; Perez-Calatayud, Jose; Simancas, Fernando; Lallena, Antonio M.; Gazdic-Santic, Maja

    2013-12-15

    Purpose: Air-communicating well ionization chambers are commonly used to assess air kerma strength of sources used in brachytherapy. The signal produced is supposed to be proportional to the air density within the chamber and, therefore, a density-independent air kerma strength is obtained when the measurement is corrected to standard atmospheric conditions using the usual temperature and pressure correction factor. Nevertheless, when assessing low energy sources, the ionization chambers may not fulfill that condition and a residual density dependence still remains after correction. In this work, the authors examined the behavior of the PTW 34051 SourceCheck ionization chamber when measuring the air kerma strength of {sup 125}I seeds.Methods: Four different SourceCheck chambers were analyzed. With each one of them, two series of measurements of the air kerma strength for {sup 125}I selectSeed{sup TM} brachytherapy sources were performed inside a pressure chamber and varying the pressure in a range from 747 to 1040 hPa (560 to 780 mm Hg). The temperature and relative humidity were kept basically constant. An analogous experiment was performed by taking measurements at different altitudes above sea level.Results: Contrary to other well-known ionization chambers, like the HDR1000 PLUS, in which the temperature-pressure correction factor overcorrects the measurements, in the SourceCheck ionization chamber they are undercorrected. At a typical atmospheric situation of 933 hPa (700 mm Hg) and 20 °C, this undercorrection turns out to be 1.5%. Corrected measurements show a residual linear dependence on the density and, as a consequence, an additional density dependent correction must be applied. The slope of this residual linear density dependence is different for each SourceCheck chamber investigated. The results obtained by taking measurements at different altitudes are compatible with those obtained with the pressure chamber.Conclusions: Variations of the altitude and

  10. Monte Carlo aided design of an improved well-type ionization chamber for low energy brachytherapy sources

    SciTech Connect

    Bohm, Tim D.; Micka, John A.; De Werd, Larry A.

    2007-04-15

    The determination of the air kerma strength of a brachytherapy seed is necessary for effective treatment planning. Well-type ionization chambers are used on site at therapy clinics to determine the air kerma strength of seeds. In this work, an improved well-type ionization chamber for low energy, low dose rate brachytherapy sources is designed using Monte Carlo transport calculations to aid in the design process. The design improvements are the elimination of the air density induced over-response effect seen in other air-communicating chambers for low energy photon sources, and a larger signal strength (response or current) for {sup 103}Pd and {sup 125}I based seeds. A prototype well chamber based on the Monte Carlo aided design but using graphite coated acrylic walls rather than the design basis air equivalent plastic (C-552) walls was constructed and experimentally evaluated. The prototype chamber produced an 85% stronger signal when measuring a commonly used {sup 103}Pd seed and a 26% stronger signal when measuring a commonly used {sup 125}I seed when compared to another commonly used well chamber. The normalized P{sub TP} corrected chamber response is, at most, 1.3% and 2.4% over unity for air densities/pressures corresponding to an elevation of 3048 m (10 000 feet) above sea level for the commonly used {sup 103}Pd and {sup 125}I based seeds respectively. Comparing calculated and measured chamber responses for common {sup 103}Pd and {sup 125}I based brachytherapy seeds show agreement within 0.6% and 0.2%, respectively. We conclude that Monte Carlo transport calculations accurately model the response of this new well chamber and in general can be used to predict the response of well chambers. The prototype chamber built in this work responds as predicted by the Monte Carlo calculations.

  11. History of the bubble chamber and related active- and internal-target nuclear tracking detectors

    NASA Astrophysics Data System (ADS)

    Becchetti, F. D.

    2015-06-01

    Donald Glaser, 1960 Nobel laureate in Physics, recently passed away (2013), as have many of his colleagues who were involved with the early development of bubble chambers at the University of Michigan. In this paper I will review those early years and the subsequent wide-spread application of active-target (AT) bubble chambers that dominated high-energy physics (HEP) research for over thirty years. Some of the related, but more modern nuclear tracking detectors being used in HEP, neutrino astrophysics and dark-matter searches also will be discussed.

  12. A TPC (Time Projection Chamber) detector for the study of high multiplicity heavy ion collisions

    SciTech Connect

    Rai, G.; Arthur, A.; Bieser, F.; Harnden, C.W.; Jones, R.; Klienfelder, S.; Lee, K.; Matis, H.S.; Nakamura, M.; McParland, C.; Nesbitt, D.; Odyniec, G.; Olson, D.; Pugh, H.G.; Ritter, H.G.; Symons, T.J.M.; Wieman, H.; Wright, M.; Wright, R. ); Rudge, A. )

    1990-01-01

    The design of a Time Projection Chamber (TPC) detector with complete pad coverage is presented. The TPC will allow the measurements of high multiplicity ({approx} 200 tracks) relativistic nucleus-nucleus collisions initiated with the heaviest, most energetic projectiles available at the LBL BEVALAC accelerator facility. The front end electronics, composed of over 15,000 time sampling channels, will be located on the chamber. The highly integrated, custom designed, electronics and the VME based data acquisition system are described. 10 refs., 8 figs., 1 tab.

  13. Scintillation properties of N2 and CF4 and performances of a scintillating ionization chamber

    NASA Astrophysics Data System (ADS)

    Lehaut, G.; Salvador, S.; Fontbonne, J.-M.; Lecolley, F.-R.; Perronnel, J.; Vandamme, Ch.

    2015-10-01

    In this work, we studied the emission yields, decay times and coincidence resolving times (CRT) of two gases, nitrogen (N2) and tetrafluoromethane (CF4), used for particle detection in the context of fission products measurement. The set-up was made of an ionization chamber and two photomultiplier tubes (PMTs) placed front-to-front on each side of the active zone of the chamber. Using the photomultiplier tubes, the number of photoelectrons (phe) converted at the photocathodes from the scintillation processes in each gas was quantified and the scintillation time spectra were recorded. A scintillation emission yield of 24 phe MeV-1 with a decay time of τd = 2.5 ns in N2, and 225 phe MeV-1 with τd = 6.2 ns for CF4, has been measured. With our set-up, the coincidence resolving time (σ values) between the two PMTs have been measured using alpha particles at 1.4 ns and 0.34 ns for N2 and CF4, respectively.

  14. Fast-neutron spectrometry using a ³He ionization chamber and digital pulse shape analysis.

    PubMed

    Chichester, D L; Johnson, J T; Seabury, E H

    2012-08-01

    Digital pulse shape analysis (dPSA) has been used with a Cuttler-Shalev type (3)He ionization chamber to measure the fast-neutron spectra of a deuterium-deuterium electronic neutron generator, a bare (252)Cf spontaneous fission neutron source, and of the transmitted fast neutron spectra of a (252)Cf source attenuated by water, graphite, liquid nitrogen, and magnesium. Rise-time dPSA has been employed using the common approach for analyzing n +(3)He→(1)H+(3)H ionization events and improved to account for wall-effect and pile-up events, increasing the fidelity of these measurements. Simulations have been performed of the different experimental arrangements and compared with the measurements, demonstrating general agreement between the dPSA-processed fast-neutron spectra and predictions. The fast-neutron resonance features of the attenuation cross sections of the attenuating materials are clearly visible within the resolution limits of the electronics used for the measurements, and the potential applications of high-resolution fast-neutron spectrometry for nuclear nonproliferation and safeguards measurements are discussed.

  15. Design and implementation of the detector control system for the BESIII drift chamber cosmic ray test

    NASA Astrophysics Data System (ADS)

    Chen, Xi-Hui; Xie, Xiao-Xi; Li, Xiao-Nan; Gao, Cui-Shan; Zhang, Yin-Hong; Nie, Zhen-Dong; Min, Jian; Xie, Yi-GANG

    2008-08-01

    After the construction of the BESIII drift chamber, a long period of cosmic rays test is necessary to verify its performance. This also provides a good opportunity to integrate the detector readout electronics and Detector Control System (DCS) into a unified working system. The goal of the DCS is to guarantee reliable physics data quality and the safe operation of the detector. It monitors and controls the HV, gas, VME crates and the environmental variables. The upper-level system is mainly developed from LabVIEW and the lower-level system mainly uses MCU and PLC technology. The system is designed to be highly flexible and scalable so that it can be applied to other detectors with little or no change. In the immediate future, it will be integrated into the entire BESIII Slow Control System.

  16. Fast transimpedance preamplifier for a boron-coated multiwire proportional chamber neutron detector

    NASA Astrophysics Data System (ADS)

    Ying, Zhang; Haiyang, Yan; Jie, Zhang; Zhijia, Sun; Ping, Cao; Qi, An; Jian, Gong

    2016-10-01

    A low-noise and fast preamplifier is needed for the new boron-coated multiwire proportional chamber (MWPC) neutron detector with a delay line readout system to improve position resolution. A transimpedance preamplifier with a rise time of 30ns, a signal-to-noise ratio higher than 40dB, and an automatic gain control function are designed to meet the aforementioned requirements. On the other hand, we also compare it to commercial preamplifiers. Then, the advantages and disadvantages of commercial and in-house preamplifiers are analyzed. The preamplifier is used and tested on a MWPC neutron detector, and results are presented.

  17. Expert System for the LHC CMS Cathode Strip Chambers (CSC) detector

    NASA Astrophysics Data System (ADS)

    Rapsevicius, Valdas; Juska, Evaldas

    2014-02-01

    Modern High Energy Physics experiments are of high demand for a generic and consolidated solution to integrate and process high frequency data streams by applying experts' knowledge and inventory configurations. In this paper we present the Expert System application that was built for the Compact Muon Solenoid (CMS) Cathode Strip Chambers (CSC) detector at the Large Hadron Collider (LHC) aiming to support the detector operations and to provide integrated monitoring. The main building blocks are the integration platform, rule-based complex event processing engine, ontology-based knowledge base, persistent storage and user interfaces for results and control.

  18. Compact cosmic ray detector for unattended atmospheric ionization monitoring.

    PubMed

    Aplin, K L; Harrison, R G

    2010-12-01

    Two vertical cosmic ray telescopes for atmospheric cosmic ray ionization event detection are compared. Counter A, designed for low power remote use, was deployed in the Welsh mountains; its event rate increased with altitude as expected from atmospheric cosmic ray absorption. Independently, Counter B's event rate was found to vary with incoming particle acceptance angle. Simultaneous co-located comparison of both telescopes exposed to atmospheric ionization showed a linear relationship between their event rates.

  19. Compact cosmic ray detector for unattended atmospheric ionization monitoring

    SciTech Connect

    Aplin, K. L.; Harrison, R. G.

    2010-12-15

    Two vertical cosmic ray telescopes for atmospheric cosmic ray ionization event detection are compared. Counter A, designed for low power remote use, was deployed in the Welsh mountains; its event rate increased with altitude as expected from atmospheric cosmic ray absorption. Independently, Counter B's event rate was found to vary with incoming particle acceptance angle. Simultaneous co-located comparison of both telescopes exposed to atmospheric ionization showed a linear relationship between their event rates.

  20. A two-dose-rate method for general recombination correction for liquid ionization chambers in pulsed beams

    NASA Astrophysics Data System (ADS)

    Tölli, Heikki; Sjögren, Rickard; Wendelsten, Mikael

    2010-08-01

    The correction for general recombination losses in liquid ionization chambers (LICs) is more complex than that in air-filled ionization chambers. The reason for this is that the saturation charge in LICs, i.e. the charge that escapes initial recombination, depends on the applied voltage. This paper presents a method, based on measurements at two different dose rates in a pulsed beam, for general recombination correction in LICs. The Boag theory for pulsed beams is used and the collection efficiency is determined by numerical methods which are equivalent to the two-voltage method used in dosimetry with air-filled ionization chambers. The method has been tested in experiments in water in a 20 MeV electron beam using two LICs filled with isooctane and tetramethylsilane. The dose per pulse in the electron beam was varied between 0.1 mGy/pulse and 8 mGy/pulse. The relative standard deviations of the collection efficiencies determined with the two-dose-rate method ranged between 0.1% and 1.5%. The dose-rate variations of the general recombination corrected charge measured with the LICs are in excellent agreement with the corresponding values obtained with an air-filled plane parallel ionization chamber.

  1. Correction factors for the INER-improved free-air ionization chambers calculated with the Monte Carlo method.

    PubMed

    Lin, Uei-Tyng; Chu, Chien-Hau

    2006-05-01

    Monte Carlo method was used to simulate the correction factors for electron loss and scattered photons for two improved cylindrical free-air ionization chambers (FACs) constructed at the Institute of Nuclear Energy Research (INER, Taiwan). The method is based on weighting correction factors for mono-energetic photons with X-ray spectra. The newly obtained correction factors for the medium-energy free-air chamber were compared with the current values, which were based on a least-squares fit to experimental data published in the NBS Handbook 64 [Wyckoff, H.O., Attix, F.H., 1969. Design of free-air ionization chambers. National Bureau Standards Handbook, No. 64. US Government Printing Office, Washington, DC, pp. 1-16; Chen, W.L., Su, S.H., Su, L.L., Hwang, W.S., 1999. Improved free-air ionization chamber for the measurement of X-rays. Metrologia 36, 19-24]. The comparison results showed the agreement between the Monte Carlo method and experimental data is within 0.22%. In addition, mono-energetic correction factors for the low-energy free-air chamber were calculated. Average correction factors were then derived for measured and theoretical X-ray spectra at 30-50 kVp. Although the measured and calculated spectra differ slightly, the resulting differences in the derived correction factors are less than 0.02%.

  2. Electrode configuration and signal subtraction technique for single polarity charge carrier sensing in ionization detectors

    DOEpatents

    Luke, P.

    1996-06-25

    An ionization detector electrode and signal subtraction apparatus and method provide at least one first conductive trace formed onto the first surface of an ionization detector. The first surface opposes a second surface of the ionization detector. At least one second conductive trace is also formed on the first surface of the ionization detector in a substantially interlaced and symmetrical pattern with the at least one first conductive trace. Both of the traces are held at a voltage potential of a first polarity type. By forming the traces in a substantially interlaced and symmetric pattern, signals generated by a charge carrier are substantially of equal strength with respect to both of the traces. The only significant difference in measured signal strength occurs when the charge carrier moves to within close proximity of the traces and is received at the collecting trace. The measured signals are then subtracted and compared to quantitatively measure the magnitude of the charge and to determine the position at which the charge carrier originated within the ionization detector. 9 figs.

  3. Electrode configuration and signal subtraction technique for single polarity charge carrier sensing in ionization detectors

    DOEpatents

    Luke, Paul

    1996-01-01

    An ionization detector electrode and signal subtraction apparatus and method provides at least one first conductive trace formed onto the first surface of an ionization detector. The first surface opposes a second surface of the ionization detector. At least one second conductive trace is also formed on the first surface of the ionization detector in a substantially interlaced and symmetrical pattern with the at least one first conductive trace. Both of the traces are held at a voltage potential of a first polarity type. By forming the traces in a substantially interlaced and symmetric pattern, signals generated by a charge carrier are substantially of equal strength with respect to both of the traces. The only significant difference in measured signal strength occurs when the charge carrier moves to within close proximity of the traces and is received at the collecting trace. The measured signals are then subtracted and compared to quantitatively measure the magnitude of the charge and to determine the position at which the charge carrier originated within the ionization detector.

  4. Investigation of electron-loss and photon scattering correction factors for FAC-IR-300 ionization chamber

    NASA Astrophysics Data System (ADS)

    Mohammadi, S. M.; Tavakoli-Anbaran, H.; Zeinali, H. Z.

    2017-02-01

    The parallel-plate free-air ionization chamber termed FAC-IR-300 was designed at the Atomic Energy Organization of Iran, AEOI. This chamber is used for low and medium X-ray dosimetry on the primary standard level. In order to evaluate the air-kerma, some correction factors such as electron-loss correction factor (ke) and photon scattering correction factor (ksc) are needed. ke factor corrects the charge loss from the collecting volume and ksc factor corrects the scattering of photons into collecting volume. In this work ke and ksc were estimated by Monte Carlo simulation. These correction factors are calculated for mono-energy photon. As a result of the simulation data, the ke and ksc values for FAC-IR-300 ionization chamber are 1.0704 and 0.9982, respectively.

  5. Determinations of the correction factors for small fields in cylindrical ionization chambers based on measurement and numerical calculation

    NASA Astrophysics Data System (ADS)

    Park, Kwangwoo; Choi, Wonhoon; Park, Sungho; Choi, Jin Hwa; Park, Suk Won; Bak, Jino

    2015-07-01

    We investigated the volume averaging effect for air-filled cylindrical ionization chambers to determine the correction factors in a small photon field for a given chamber. We measured output factors with several cylindrical ionization chambers, and by using a mathematical method similar to deconvolution, we modeled the non-constant and inhomogeneous exposure function in the cavity of the chamber. The parameters in the exposure function and the correction factors were determined by solving a system of equations that we had developed by using the measured data and the geometry of the given chamber. The correction factors (CFs) were very similar to those obtained from Monte Carlo (MC) simulations. For example, the CFs in this study were found to be 1.116 for PTW31010 and 1.0225 for PTW31016 while the CFs obtained from MC simulations were reported as being approximately between 1.17 and 1.20 for PTW31010 and between 1.02 and 1.06 for PTW31016 in a 6-MV photon beam of 1 × 1 cm2. Furthermore, the method of deconvolution combined with the MC result for the chamber's response function showed a similar CF for PTW 30013, which was reported as 2.29 and 1.54 for a 1 × 1 cm2 and a 1.5 × 1.5 cm2 field size, respectively. The CFs from our method were similar, 2.42 and 1.54. In addition, we report CFs for PTW30013, PTW31010, PTW31016, IBA FC23-C, and IBA CC13. As a consequence, we suggest the use of our method to measure the correct output factor by using the fact that an inhomogeneous exposure causes a volume averaging effect in the cavity of air-filled cylindrical ionization chamber. The result obtained by using our method is very similar to that obtained from MC simulations. The method we developed can easily be applied in clinics.

  6. Comparison of dose measurements in CT using a novel semiconductor detector and a small ion chamber

    SciTech Connect

    Paschoal, Cinthia M. M.; Ferreira, Fernanda Carla L.; Santos, Luiz A. P.; Souza, Divanizia N.

    2015-07-01

    The advance of multislice computed tomography (CT) has become inadequate the currently dosimetric protocol used in CT. Instead of dosimetry based on the measurement of CTDI using a pencil ion chamber of 100 m of length, it was proposed the use of a small ion chamber (IC) and the calculating the dose equilibrium (Deq) at the location of the chamber. The objective of this work was to compare the performance of a short IC and a commercial photodiode to measure the accumulated dose at the center of the scan length L, DL(0), and to obtain the equilibrium dose Deq using the two detectors. The result for L=100 mm was compared with the result of a pencil chamber. The results indicate that the commercial photodiode is suitable to measure the accumulated dose at the center of the scan length L as compared with the ion chambers. This methodology allows measurements of the accumulated dose for any desired scan length, allowing measuring the equilibrium dose Deq if the phantom is long enough to allow it. (authors)

  7. Front-End electronics development for the new Resistive Plate Chamber detector of HADES

    NASA Astrophysics Data System (ADS)

    Gil, A.; Belver, D.; Cabanelas, P.; Díaz, J.; Garzón, J. A.; González-Díaz, D.; Koenig, W.; Lange, J. S.; Marín, J.; Montes, N.; Skott, P.; Traxler, M.

    2007-11-01

    In this paper we present the new RPC wall, which is being installed in the HADES detector at Darmstadt GSI. It consists of time-of-flight (TOF) detectors used for both particle identification and triggering. Resistive Plate Chamber (RPC) detectors are becoming widely used because of their excellent TOF capabilities and reduced cost. The wall will contain 1024 RPC modules, covering an active area of around 7 m2, replacing the old TOFino detector at the low polar angle region. The excellent TOF and good charge resolutions of the new detector will improve the time resolution to values better than 100 ps. The Front-End electronics for the readout of the RPC signals is implemented with two types of boards to satisfy the space constraints: the Daughterboards are small boards that amplify the low level signals from the detector and provide fast discriminators for time of flight measurements, as well as an integrator for charge measurements. The Motherboard provides stable DC voltages and a stable ground, threshold DACs for the discriminators, multiplicity trigger and impedance matched paths for transfer of time window signals that contain information about time and charge. These signals are sent to a custom TDC board that label each event and send data through Ethernet to be conveniently stored.

  8. High resolution resonance ionization imaging detector and method

    DOEpatents

    Winefordner, James D.; Matveev, Oleg I.; Smith, Benjamin W.

    1999-01-01

    A resonance ionization imaging device (RIID) and method for imaging objects using the RIID are provided, the RIID system including a RIID cell containing an ionizable vapor including monoisotopic atoms or molecules, the cell being positioned to intercept scattered radiation of a resonance wavelength .lambda..sub.1 from the object which is to be detected or imaged, a laser source disposed to illuminate the RIID cell with laser radiation having a wavelength .lambda..sub.2 or wavelengths .lambda..sub.2, .lambda..sub.3 selected to ionize atoms in the cell that are in an excited state by virtue of having absorbed the scattered resonance laser radiation, and a luminescent screen at the back surface of the RIID cell which presents an image of the number and position of charged particles present in the RIID cell as a result of the ionization of the excited state atoms. The method of the invention further includes the step of initially illuminating the object to be detected or imaged with a laser having a wavelength selected such that the object will scatter laser radiation having the resonance wavelength .lambda..sub.1.

  9. Liquid ionization chamber initial recombination dependence on LET for electrons and photons.

    PubMed

    Johansson, Erik; Andersson, Jonas; Johansson, Lennart; Tölli, Heikki

    2013-06-21

    The possibility of indirect measurements of linear energy transfer (LET) with a liquid ionization chamber (LIC) has been investigated by studying initial recombination losses at different applied voltages. A linear fit is made to the voltage-signal curve and the intersection point of the fit and the voltage-axis is shown to correlate with LET. The LIC applied voltages were 100-700 V, which corresponds to electric field strengths between 0.3 and 2.0 MV m(-1). Several different photon and electron beams have been studied, and by using MCNPX™ the respective LET spectra have been determined. The beam qualities in this study were found to have a fluence averaged LET between 0.17 and 1.67 keV µm(-1) and a corresponding dose averaged LET between 0.97 and 4.62 keV µm(-1). For the experimental data in this study the linear fit method yields consistent results with respect to Monte Carlo simulated LET values. A calibration curve for LET determination is provided for the LIC used in the present work.

  10. Extended virtual detector theory for strong-field atomic ionization.

    PubMed

    Wang, Xu; Tian, Justin; Eberly, J H

    2013-06-14

    For time-dependent strong-field atomic ionization a new theoretical approach is described that combines the numerical time-dependent Schrödinger equation (TDSE) and the numerical time-dependent Newtonian equation (TDNE). This approach keeps both the accuracy of quantum calculations and the speed of classical calculations. It does not use approximate tunneling formulas. It is applied to a recent experimental result, and we show its successful comparison to extensive TDSE calculations made under exactly the same conditions.

  11. DEPOSITS ON FLAME IONIZATION DETECTORS WITH SILICONE GUM RUBBER COLUMNS,

    DTIC Science & Technology

    During the preparative gas chromatography of some organophosphorus esters using a silicone gum rubber column, a copious white deposit was formed on...detector deposits were mainly silicon phosphate (2SiO2.P2O5), and the injection port samples were alpha cristobalite (SiO2).

  12. The U.S. experience with smoke detectors: who has them? How well do they work? When don't they work?

    PubMed

    Hall, J R

    1994-01-01

    From 1975 to 1984, the United States experienced remarkable growth in the use of home smoke detectors, making the home smoke detector the fire safety success story of the decade. As of 1993, only 1 home in 12 remained unprotected.

  13. Acoustic studies for alpha background rejection in dark matter bubble chamber detectors

    SciTech Connect

    Bou-Cabo, M.; Felis, I.; Ardid, M.; Collaboration: COUPP Collaboration

    2013-08-08

    COUPP (Chicagoland Observatory for Underground Particle Physics) is an experiment with bubble chambers able to detect dark matter directly either with Spin-Dependent or with Spin-Independent interactions. The target material is a superheated liquid (usually CF3I) that can be bubble nucleated due to nuclear recoils produced by elastic collisions of dark matter particles. The bubble growth inside the chamber is accompanied with an acoustic signature. The acoustic technique has been successfully used to have a good alpha discrimination (about 99%). In this paper, we present different studies and results related with the characterization of the acoustic properties of the detector and the different phenomena involved in the acoustic measurements of the bubble growth, such as sound generation, sound transmission and optimization of piezoelectric transducers.

  14. SU-E-T-414: Experimental Correction of High-Z Electrode Effect in Mini-Ionization Chambers for Small Beam Dosimetry

    SciTech Connect

    Larraga-Gutierrez, J

    2015-06-15

    Purpose: To correct for the over-response of mini-ionization chambers with high-Z central electrodes. The hypothesis is that by applying a negative/reverse voltage, it is possible to suppress the signal generated in the high-Z central electrode by low-energy photons. Methods: The mini-ionization chambers used in the experiments were a PTW-31014, PTW-31006 and IBA-CC01. The PTW-31014 has an aluminum central electrode while the PTW-31006 and IBA-CC01 have a steel one. Total scatter factors (Scp) were measured for a 6 MV photon beam down to a square field size of 0.5 cm. The measurements were performed in water at 10 cm depth with SAD of 100 cm. The Scp were measured with the dosimeters with +400V bias voltage. In the case of the PTW-31006 and IBA-CC01, the measurements were repeated with −400V bias voltage. Also, the field factors in water were calculated with Monte Carlo simulations for comparison. Results: The measured Scp at +400V with the PTW-31006 and IBA-CC01 detectors were in agreement within 0.2% down to a field size of 1.5 cm. Both dosimeters shown a systematic difference about 2.5% with the Scp measured with the PTW-31014 and the Monte Carlo calculated field factors. The measured Scp at −400V with the PTW-31006 and IBA-CC01 detectors were in close agreement with the PTW-31014 measured Scp and the field factors within 0.3 and 1.0%, respectively. In the case of the IBA-CC01 it was found a good agreement (1%) down to field size of 1.0 cm. All the dosimeters shown differences up to 17% between the measured Scp and the field factor for the 0.5 cm field size. Conclusion: By applying a negative/reverse voltage to the mini-ionization chambers with high-Z central electrode it was possible to correct for their over-response to low energy photons.

  15. Polarization of silicon detectors by minimum ionizing particles

    NASA Astrophysics Data System (ADS)

    Dezillie, B.; Eremin, V.; Li, Z.; Verbitskaya, E.

    2000-10-01

    This work presents quantitative predictions of the properties of highly irradiated (e.g. by high-energy particles, up to an equivalent fluence of 1×10 14 n cm -2) silicon detectors operating at cryogenic temperature. It is shown that the exposure to the Minimum Ionising Particle (MIP) with counting rates of about 10 6 cm -2 s -1 can influence the electric field distribution in the detector's sensitive volume. This change in the electric field distribution and its effect on the charge collection efficiency are discussed in the frame of a model based on trapping of carriers generated by MIPs. The experiment was performed at 87 K with an infrared (1030 nm) laser to simulate MIPs.

  16. Determination of the thermal neutron flux in a fast neutron beam by use of a boron-coated ionization chamber.

    PubMed

    Lüdemann, L; Matzen, T; Matzke, M; Schmidt, R; Scobel, W

    1995-11-01

    The thermal neutron distribution in slow and fast neutron beams is usually determined using the foil activation method. In this work a small magnesium walled ionization chamber, in which the inner surface of the wall has been coated with 10B to increase the sensitivity for thermal neutrons, is used to estimate the thermal neutron component of the beam. After calibration and determination of the directional response in a thermal neutron beam a comparison with foil activation at different depths in water was performed to investigate the reliability of the ionization measurements. The chamber was used in a computer controlled water phantom to measure the depth and lateral distribution of the thermal neutron dose. With this arrangement two-dimensional scans of the thermal neutrons could be performed quickly and with high accuracy.

  17. Using Ionizing Radiation Detectors. Module 11. Vocational Education Training in Environmental Health Sciences.

    ERIC Educational Resources Information Center

    Consumer Dynamics Inc., Rockville, MD.

    This module, one of 25 on vocational education training for careers in environmental health occupations, contains self-instructional materials on using ionizing radiation detectors. Following guidelines for students and instructors and an introduction that explains what the student will learn are three lessons: (1) naming and telling the function…

  18. 40 CFR 1065.267 - Gas chromatograph with a flame ionization detector.

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... 40 Protection of Environment 33 2014-07-01 2014-07-01 false Gas chromatograph with a flame ionization detector. 1065.267 Section 1065.267 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR POLLUTION CONTROLS ENGINE-TESTING PROCEDURES Measurement Instruments Hydrocarbon...

  19. 40 CFR 1065.267 - Gas chromatograph with a flame ionization detector.

    Code of Federal Regulations, 2013 CFR

    2013-07-01

    ... 40 Protection of Environment 34 2013-07-01 2013-07-01 false Gas chromatograph with a flame ionization detector. 1065.267 Section 1065.267 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR POLLUTION CONTROLS ENGINE-TESTING PROCEDURES Measurement Instruments Hydrocarbon...

  20. 40 CFR 1065.267 - Gas chromatograph with a flame ionization detector.

    Code of Federal Regulations, 2012 CFR

    2012-07-01

    ... 40 Protection of Environment 34 2012-07-01 2012-07-01 false Gas chromatograph with a flame ionization detector. 1065.267 Section 1065.267 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR POLLUTION CONTROLS ENGINE-TESTING PROCEDURES Measurement Instruments Hydrocarbon...

  1. A Bubble Chamber Revival: Superheated Liquid Detectors for Dark Matter Searches and Other Applications

    NASA Astrophysics Data System (ADS)

    Sonnenschein, Andrew

    2012-10-01

    Bubble chambers were recently deemed to be obsolete, of interest only to historians, but a few new applications have unexpectedly emerged in nuclear and particle astrophysics. These include the search for WIMP dark matter and the measurement of a few otherwise intractable nuclear cross sections. The new bubble chambers exploit features of the bubble nucleation process that were unappreciated or irrelevant in the 1950s-1970s when the technology was known for its capability to provide fine-grained tracking of high energy particles in a dense target medium. By carefully tuning the temperature and pressure, a liquid can be made selectively sensitive to particles on the basis of their specific rate of energy loss, enabling a high degree of background rejection power when searching for rare heavily-ionizing tracks. Surprisingly, additional information on the microstructure of particle tracks can be extracted from the acoustic noise produce by bubble nucleation. Other novel features of the new bubble chambers include the use of digital photography, self-triggering, and the achievement of nearly continuous sensitivity by the avoidance of bubble nucleation on internal wetted surfaces.

  2. Development of a Bioaerosol single particle detector (BIO IN) for the Fast Ice Nucleus CHamber FINCH

    NASA Astrophysics Data System (ADS)

    Bundke, U.; Reimann, B.; Nillius, B.; Jaenicke, R.; Bingemer, H.

    2010-02-01

    In this work we present the setup and first tests of our new BIO IN detector. This detector was constructed to classify atmospheric ice nuclei (IN) for their biological content. It is designed to be coupled to the Fast Ice Nucleus CHamber FINCH. If one particle acts as an ice nucleus, it will be at least partly covered with ice at the end of the development section of the FINCH chamber. The device combines an auto-fluorescence detector and a circular depolarization detector for simultaneous detection of biological material and discrimination between water droplets, ice crystals and non activated large aerosol particles. The excitation of biological material with UV light and analysis of auto-fluorescence is a common principle used for flow cytometry, fluorescence microscopy, spectroscopy and imaging. The detection of auto-fluorescence of airborne single particles demands some more experimental effort. However, expensive commercial sensors are available for special purposes, e.g. size distribution measurements. But these sensors will not fit the specifications needed for the FINCH IN counter (e.g. high sample flow of up 10 LPM). The newly developed -low cost- BIO IN sensor uses a single high-power UV LED for the electronic excitation instead of much more expensive UV lasers. Other key advantages of the new sensor are the low weight, compact size, and the little effect on the aerosol sample, which allows it to be coupled with other instruments for further analysis. The instrument will be flown on one of the first missions of the new German research aircraft "HALO" (High Altitude and LOng range).

  3. SU-E-T-377: Evaluation of a Novel Transmission Detector as a Reference Chamber for Use in Beam Measurements

    SciTech Connect

    Epstein, D; Shekel, E; Levin, D

    2015-06-15

    Purpose: To evaluate a new transmission chamber for use as a reference chamber in measurements of beam data Methods: We assessed the performance of a new transmission detector, the Stealth Chamber, manufactured by IBA (IBA Dosimetry). The chamber has an active volume of 249 cm{sup 3}, with an attenuation equivalent of <0.5mm Al. We mounted the chamber to a TrueBeam linac (Varian Medical Systems, Palo Alto, CA) such that the active area is perpendicular to the beam direction. We performed PDD and profile measurements on field sizes from 1×1 cm{sup 2} to 10×10 cm{sup 2}, as well as for a 4 mm cone. The field detector was either a CC-13 chamber (IBA Dosimetry) with an active volume of 0.13 cm{sup 3} or an Edge Detector (Sun Nuclear, Melbourne, FL) with an active volume of 0.019 cm{sup 3}. For comparison we repeated all measurements using a CC-13 or CC-01 (active volume 0.01 cm{sup 3} ) chamber as reference detectors (IBA Dosimetry).All scans were acquired using a Blue Phantom2 (IBA Dosimetry) and IBA OmniPro-Accept v7.4.24 software. Results: All measurements with the Stealth Chamber were identical to those with ion reference chambers. For both regular and filter free 6MV beams there was agreement in PDDs and profiles, including the penumbra region, for all field sizes. This was true for the 4mm cone measurement, as well. The deviation between the Stealth and ion chamber measurements was on average 0.3%. Conclusion: The Stealth Chamber gives identical beam data as a conventional ion chamber for all field sizes. The advantage of the Stealth chamber over ion chambers is its efficiency. Once mounted, there is no need to reposition the chamber with varying field sizes. This translates into a huge savings in measurement time, as well as a reduction in potential errors due to reference chamber mispositioning.

  4. The stability of liquid-filled matrix ionization chamber electronic portal imaging devices for dosimetry purposes.

    PubMed

    Louwe, R J W; Tielenburg, R; van Ingen, K M; Mijnheer, B J; van Herk, M B

    2004-04-01

    This study was performed to determine the stability of liquid-filled matrix ionization chamber (LiFi-type) electronic portal imaging devices (EPID) for dosimetric purposes. The short- and long-term stability of the response was investigated, as well as the importance of factors influencing the response (e.g., temperature fluctuations, radiation damage, and the performance of the electronic hardware). It was shown that testing the performance of the electronic hardware as well as the short-term stability of the imagers may reveal the cause of a poor long-term stability of the imager response. In addition, the short-term stability was measured to verify the validity of the fitted dose-response curve immediately after beam startup. The long-term stability of these imagers could be considerably improved by correcting for room temperature fluctuations and gradual changes in response due to radiation damage. As a result, the reproducibility was better than 1% (1 SD) over a period of two years. The results of this study were used to formulate recommendations for a quality control program for portal dosimetry. The effect of such a program was assessed by comparing the results of portal dosimetry and in vivo dosimetry using diodes during the treatment of 31 prostate patients. The improvement of the results for portal dosimetry was consistent with the deviations observed with the reproducibility tests in that particular period. After a correction for the variation in response of the imager, the average difference between the measured and prescribed dose during the treatment of prostate patients was -0.7%+/-1.5% (1 SD), and -0.6%+/-1.1% (1 SD) for EPID and diode in vivo dosimetry, respectively. It can be concluded that a high stability of the response can be achieved for this type of EPID by applying a rigorous quality control program.

  5. SMOKE: Characterization of Smoke Particulate for Spacecraft Fire Detection

    NASA Technical Reports Server (NTRS)

    Urban, D. L.; Mulholland, G.; Yuan, Z. G.; Yang, J.; Cleary, T.

    2001-01-01

    'Smoke' is a flight definition investigation whose purpose is to characterize the smoke particulate from microgravity smoke sources to enable improved design of future space-craft smoke detectors. In the earliest missions (Mercury, Gemini and Apollo), the crew quarters were so cramped that it was considered reasonable that the astronauts would rapidly detect any fire. The Skylab module, however, included approximately 30 UV-sensing fire detectors. The Space Shuttle Orbiter has nine particle-ionization smoke detectors in the mid-deck and flight deck. The detectors for the US segments of the International Space Station (ISS) are laser-diode, forward-scattering, smoke detectors. Current plans for the ISS call for two detectors in the open area of the module, and detectors in racks that have cooling air-flow. Due to the complete absence of microgravity data, all three of these detector systems were designed based upon 1-g test data and experience. As planned mission durations and complexity increase and the volume of spacecraft increases, the need for and importance of effective, crew-independent, fire detection will grow significantly, necessitating more research into microgravity fire phenomena. In 1997 the Comparative Soot Diagnostics Experiment (CSD) flew in the Orbiter Middeck as a Glovebox payload. The CSD experiment was designed to produce small quantities of smoke from several sources to obtain particulate samples and to determine the response of the ISS and Orbiter smoke detectors to these sources. Marked differences in the performance of the detectors compared to their behavior in 1-g were observed. In extreme cases, the detector used in the orbiter was completely blind to easily visible smoke from sources that were readily detected in 1-g. It is hypothesized but as yet unverified that this performance difference was due to enhanced growth of liquid smoke droplets in low-g. These CSD results clearly demonstrate that spacecraft smoke detector design cannot be

  6. Predicting the yield performances of multicomponent oil smokes confined in closed chambers. Technical report Jan-Nov 80

    SciTech Connect

    Rubel, G.O.

    1983-06-01

    A semiempirical model is developed to predict the yield performances of multicomponent oil smokes confined in closed chambers. The model employs vapor pressure-percent mass evaporated correlations, which were derived from oil droplet evaporation studies, to predict the saturation requirements of fog, oil, PEG 200 and No. 2 diesel fuel. Comparison of model predictions and experiment indicates good agreement for fog oil and PEG 200 vaporization/condensation yield tests. Both model and experiment demonstrate the inferior yield performance of No. 2 diesel fuel, possessing a yield less than half that of either fog oil or PEG 200. The lower yield performance of No. 2 diesel fuel is due to the high volatility of diesel fuel as compared to fog oil or PEG 200. The model also illustrates the dependency of multicomponent oil yield performances on chamber volume. As a result, while fog oil and PEG200 will be characterized by similar yields at a specific chamber volume, the yield equivalency will not hold at larger chamber volumes. Implications on the liquid smoke-open atmosphere persistency characteristics are discussed. Finally, the hygroscopic nature of PEG 200 is discussed and a model is developed for the prediction of the yield performance of PEG 200 as a function of relative humidity. It is also shown the yield of PEG 200 can exceed that of fog oil under high-relative humidity conditions.

  7. Direct measurement of absorbed dose to water in HDR {sup 192}Ir brachytherapy: Water calorimetry, ionization chamber, Gafchromic film, and TG-43

    SciTech Connect

    Sarfehnia, Arman; Kawrakow, Iwan; Seuntjens, Jan

    2010-04-15

    Purpose: Gafchromic film and ionometric calibration procedures for HDR {sup 192}Ir brachytherapy sources in terms of dose rate to water are presented and the experimental results are compared to the TG-43 protocol as well as with the absolute dose measurement results from a water calorimetry-based primary standard. Methods: EBT-1 Gafchromic films, an A1SL Exradin miniature Shonka thimble type chamber, and an SI HDR 1000 Plus well-type chamber (Standard Imaging, Inc., Middleton, WI) with an ADCL traceable S{sub k} calibration coefficient (following the AAPM TG-43 protocol) were used. The Farmer chamber and Gafchromic film measurements were performed directly in water. All results were compared to direct and absolute absorbed dose to water measurements from a 4 deg. C stagnant water calorimeter. Results: Based on water calorimetry, the authors measured the dose rate to water to be 361{+-}7 {mu}Gy/(h U) at a 55 mm source-to-detector separation. The dose rate normalized to air-kerma strength for all the techniques agree with the water calorimetry results to within 0.83%. The overall 1-sigma uncertainty on water calorimetry, ionization chamber, Gafchromic film, and TG-43 dose rate measurement amounts to 1.90%, 1.44%, 1.78%, and 2.50%, respectively. Conclusions: This work allows us to build a more realistic uncertainty estimate for absorbed dose to water determination using the TG-43 protocol. Furthermore, it provides the framework necessary for a shift from indirect HDR {sup 192}Ir brachytherapy dosimetry to a more accurate, direct, and absolute measurement of absorbed dose to water.

  8. A sensitive gas chromatography detector based on atmospheric pressure chemical ionization by a dielectric barrier discharge.

    PubMed

    Kirk, Ansgar T; Last, Torben; Zimmermann, Stefan

    2017-02-03

    In this work, we present a novel concept for a gas chromatography detector utilizing an atmospheric pressure chemical ionization which is initialized by a dielectric barrier discharge. In general, such a detector can be simple and low-cost, while achieving extremely good limits of detection. However, it is non-selective apart from the use of chemical dopants. Here, a demonstrator manufactured entirely from fused silica capillaries and printed circuit boards is shown. It has a size of 75×60×25mm(3) and utilizes only 2W of power in total. Unlike other known discharge detectors, which require high-purity helium, this detector can theoretically be operated using any gas able to form stable ion species. Here, purified air is used. With this setup, limits of detection in the low parts-per-billion range have been obtained for acetone.

  9. Experimental determination of the effective point of measurement of cylindrical ionization chambers for high-energy photon and electron beams.

    PubMed

    Huang, Yanxiao; Willomitzer, Christian; Zakaria, Golam Abu; Hartmann, Guenther H

    2010-01-01

    Measurements of depth-dose curves in water phantom using a cylindrical ionization chamber require that its effective point of measurement is located at the measuring depth. Recommendations for the position of the effective point of measurement with respect to the central axis valid for high-energy electron and photon beams are given in dosimetry protocols. According to these protocols, the use of a constant shift P(eff) is currently recommended. However, this is still based on a very limited set of experimental results. It is therefore expected that an improved knowledge of the exact position of the effective point of measurement will further improve the accuracy of dosimetry. Recent publications have revealed that the position of the effective point of measurement is indeed varying with beam energy, field size and also with chamber geometry. The aim of this study is to investigate whether the shift of P(eff) can be taken to be constant and independent from the beam energy. An experimental determination of the effective point of measurement is presented based on a comparison between cylindrical chambers and a plane-parallel chamber using conventional dosimetry equipment. For electron beams, the determination is based on the comparison of halfvalue depth R(50) between the cylindrical chamber of interest and a well guarded plane-parallel Roos chamber. For photon beams, the depth of dose maximum, d(max), the depth of 80% dose, d(80), and the dose parameter PDD(10) were used. It was again found that the effective point of measurement for both, electron and photon beams Dosimetry, depends on the beam energy. The deviation from a constant value remains very small for photons, whereas significant deviations were found for electrons. It is therefore concluded that use of a single upstream shift value from the centre of the cylindrical chamber as recommended in current dosimetry protocols is adequate for photons, however inadequate for accurate electron beam dosimetry.

  10. Characterization of radiation beams used to determinate the correction factor for a CyberKnife® unit reference field using ionization chambers

    SciTech Connect

    Aragón-Martínez, Nestor Massillon-JL, Guerda; Gómez-Muñoz, Arnulfo

    2014-11-07

    This paper aimed to characterize a 6 MV x-ray beam from a Varian® iX linear accelerator in order to obtain the correction factors needed by the IAEA/AAPM new formalism{sup 1}. The experiments were performed in a liquid water phantom under different irradiation conditions: a) Calibration of the reference field of 10 cm × 10 cm at 90 cm SSD and 10 cm depth was carried out according to the TRS-398 protocol using three ionization chambers (IC) calibrated in different reference laboratory and b) Measurement of the absorbed dose rate at 70 cm SSD and 10 cm depth in a 10 cm × 10 cm and 5.4 cm × 5.4 cm fields was obtained in order to simulate the CyberKnife® conditions where maximum distance between the source and the detector is equal to 80 cm and the maximum field size is 6 cm diameter. Depending where the IC was calibrated, differences between 0.16% and 2.24% in the absorbed dose rate measured in the 10 cm × 10 cm field at 90 cm SSD were observed, while for the measurements at 70 cm SSD, differences between 1.27% and 3.88% were obtained. For the 5.4 cm × 5.4 cm field, the absorbed dose measured with the three ICs varies between 1.37% and 3.52%. The increase in the difference on the absorbed dose when decreasing the SSD could possibly be associated to scattering radiation generated from the collimators and/or the energy dependence of the ionization chambers to low-energy radiation. The results presented in this work suggest the importance of simulating the CyberKnife® conditions using other linear accelerator for obtaining the correction factors as proposed by the IAEA/AAPM new formalism in order to measure the absorbed dose with acceptable accuracy.

  11. Characterization of radiation beams used to determinate the correction factor for a CyberKnife® unit reference field using ionization chambers

    NASA Astrophysics Data System (ADS)

    Aragón-Martínez, Nestor; Gómez-Muñoz, Arnulfo; Massillon-JL, Guerda

    2014-11-01

    This paper aimed to characterize a 6 MV x-ray beam from a Varian® iX linear accelerator in order to obtain the correction factors needed by the IAEA/AAPM new formalism1. The experiments were performed in a liquid water phantom under different irradiation conditions: a) Calibration of the reference field of 10 cm × 10 cm at 90 cm SSD and 10 cm depth was carried out according to the TRS-398 protocol using three ionization chambers (IC) calibrated in different reference laboratory and b) Measurement of the absorbed dose rate at 70 cm SSD and 10 cm depth in a 10 cm × 10 cm and 5.4 cm × 5.4 cm fields was obtained in order to simulate the CyberKnife® conditions where maximum distance between the source and the detector is equal to 80 cm and the maximum field size is 6 cm diameter. Depending where the IC was calibrated, differences between 0.16% and 2.24% in the absorbed dose rate measured in the 10 cm × 10 cm field at 90 cm SSD were observed, while for the measurements at 70 cm SSD, differences between 1.27% and 3.88% were obtained. For the 5.4 cm × 5.4 cm field, the absorbed dose measured with the three ICs varies between 1.37% and 3.52%. The increase in the difference on the absorbed dose when decreasing the SSD could possibly be associated to scattering radiation generated from the collimators and/or the energy dependence of the ionization chambers to low-energy radiation. The results presented in this work suggest the importance of simulating the CyberKnife® conditions using other linear accelerator for obtaining the correction factors as proposed by the IAEA/AAPM new formalism in order to measure the absorbed dose with acceptable accuracy.

  12. Investigation of particle and vapor wall-loss effects on controlled wood-smoke smog-chamber experiments

    NASA Astrophysics Data System (ADS)

    Bian, Q.; May, A. A.; Kreidenweis, S. M.; Pierce, J. R.

    2015-06-01

    Smog chambers are extensively used to study processes that drive gas and particle evolution in the atmosphere. A limitation of these experiments is that particles and gas-phase species may be lost to chamber walls on shorter timescales than the timescales of the atmospheric processes being studied in the chamber experiments. These particle and vapor wall losses have been investigated in recent studies of secondary organic aerosol (SOA) formation, but they have not been systematically investigated in experiments of primary emissions from combustion. The semi-volatile nature of combustion emissions (e.g. from wood smoke) may complicate the behavior of particle and vapor wall deposition in the chamber over the course of the experiments due to the competition between gas/particle and gas/wall partitioning. Losses of vapors to the walls may impact particle evaporation in these experiments, and potential precursors for SOA formation from combustion may be lost to the walls, causing underestimates of aerosol yields. Here, we conduct simulations to determine how particle and gas-phase wall losses contributed to the observed evolution of the aerosol during experiments in the third Fire Lab At Missoula Experiment (FLAME III). We use the TwO-Moment Aerosol Sectional (TOMAS) microphysics algorithm coupled with the organic volatility basis set (VBS) and wall-loss formulations to examine the predicted extent of particle and vapor wall losses. We limit the scope of our study to the dark periods in the chamber before photo-oxidation to simplify the aerosol system for this initial study. Our model simulations suggest that over one third of the initial particle-phase organic mass (36%) was lost during the experiments, and roughly half of this particle organic mass loss was from direct particle wall loss (56% of the loss) with the remainder from evaporation of the particles driven by vapor losses to the walls (44% of the loss). We perform a series of sensitivity tests to understand

  13. Investigation of particle and vapor wall-loss effects on controlled wood-smoke smog-chamber experiments

    NASA Astrophysics Data System (ADS)

    Bian, Q.; May, A. A.; Kreidenweis, S. M.; Pierce, J. R.

    2015-10-01

    Smog chambers are extensively used to study processes that drive gas and particle evolution in the atmosphere. A limitation of these experiments is that particles and gas-phase species may be lost to chamber walls on shorter timescales than the timescales of the atmospheric processes being studied in the chamber experiments. These particle and vapor wall losses have been investigated in recent studies of secondary organic aerosol (SOA) formation, but they have not been systematically investigated in experiments of primary emissions from combustion. The semi-volatile nature of combustion emissions (e.g. from wood smoke) may complicate the behavior of particle and vapor wall deposition in the chamber over the course of the experiments due to the competition between gas/particle and gas/wall partitioning. Losses of vapors to the walls may impact particle evaporation in these experiments, and potential precursors for SOA formation from combustion may be lost to the walls, causing underestimations of aerosol yields. Here, we conduct simulations to determine how particle and gas-phase wall losses contributed to the observed evolution of the aerosol during experiments in the third Fire Lab At Missoula Experiment (FLAME III). We use the TwO-Moment Aerosol Sectional (TOMAS) microphysics algorithm coupled with the organic volatility basis set (VBS) and wall-loss formulations to examine the predicted extent of particle and vapor wall losses. We limit the scope of our study to the dark periods in the chamber before photo-oxidation to simplify the aerosol system for this initial study. Our model simulations suggest that over one-third of the initial particle-phase organic mass (41 %) was lost during the experiments, and over half of this particle-organic mass loss was from direct particle wall loss (65 % of the loss) with the remainder from evaporation of the particles driven by vapor losses to the walls (35 % of the loss). We perform a series of sensitivity tests to understand

  14. Characterization of 2 MeV, 4 MeV, 6 MeV and 18 MeV buildup caps for use with a 0.6 cubic centimeter thimble ionization chamber

    SciTech Connect

    Salyer, R.L.; VanDenburg, J.W.; Prinja, A.K.; Kirby, T.; Busch, R.; Hong-Nian Jow

    1996-07-01

    The purpose of this research is to characterize existing 2 MeV, 4 MeV and 6 MeV buildup caps, and to determine if a buildup cap can be made for the 0.6 cm{sup 3} thimble ionization chamber that will accurately measure exposures in a high-energy photon radiation field. Two different radiation transport codes were used to computationally characterize existing 2 MeV, 4 MeV, and 6 MeV buildup caps for a 0.6 cm{sup 3} active volume thimble ionization chamber: ITS, The Integrated TIGER Series of Coupled Electron-Photon Monte Carlo Transport Codes; and CEPXS/ONEDANT, A One-Dimensional Coupled Electron-Photon Discrete Ordinates Code Package. These codes were also used to determine the design characteristics of a buildup cap for use in the 18 MeV photon beam produced by the 14 TW pulsed power HERMES-III electron accelerator. The maximum range of the secondary electron, the depth at which maximum dose occurs, and the point where dose and collision kerma are equal have been determined to establish the validity of electronic equilibrium. The ionization chamber with the appropriate buildup cap was then subjected to a 4 MeV and a 6 MeV bremmstrahlung radiation spectrum to determine the detector response.

  15. Development of TOF-PET detectors based on the Multi-Gap Resistive Plate Chambers

    NASA Astrophysics Data System (ADS)

    Amaldi, U.; Borghi, G.; Bucciantonio, M.; Kieffer, R.; Samarati, J.; Sauli, F.; Watts, D.

    2015-04-01

    We describe the development, construction and preliminary results obtained with medium-size Multi-Gap Resistive Plate Chambers prototypes designed to detect and localize 511 keV photons for Positron Emission Tomography imaging applications. The devices are intended for in-beam monitoring of the treatment plans throughout deep tumor therapy with hadron beams; emphasis is put on achieving coincidence time resolutions of few hundred ps, in order to exploit optimized reconstruction algorithm and reduce the heavy non-correlated background contributions distinctive of this operation. Using technologies developed for high energy physics experiments, the detectors can be built for covering large areas, thus leading the way to the conception of full-body PET systems at low cost.

  16. Patient specific quality assurance of RapidArc pre treatment plans using semiflex 0.125 cc ionization chamber

    NASA Astrophysics Data System (ADS)

    Kumar, S. A. Syam; Vivekanandan, Nagarajan

    2017-01-01

    Patient specific pre-treatment quality assurance for RapidArc plans were analyzed for hundred patients for different sites. Verification plan was created for each treatment plan in Eclipse 8.6 treatment planning system with the semiflex ionization chamber and the octavius phantom. Absolute point dose were measured for head and neck, thorax and abdomen cases using semiflex (0.125 cc) ionization chamber. Positive absolute mean dose variation of 0.56% was observed with thorax cases with a standard deviation (SD) of ±1.13 between the plans with a range of -1.78% to 2.70%. Negative percentage dose errors were found with head and neck and abdomen cases, with a mean variation of -0.43% (SD±1.50), (range -3.25% to 2.85%) and -0.35% (SD±1.48), (range -3.10% to 2.65%) for head and neck and abdomen cases respectively. Evaluation has been done using PTW verisoft software by keeping the passing criteria as 3 mm DTA, 3% DD, for 95% of the evaluated dose points. The maximum percentage value failed in gamma analysis was found to be 4.95, 4.75, and 4.88 for head and neck, thorax, and abdomen cases respectively. In all the cases analyzed the percentage dose points failed the gamma criteria is less than 5%. On the basis of the studies performed it can be concluded that the semiflex ionization chamber having a volume of 0.125 cc can be used efficiently for measuring the pre-treatment quality assurance of RapidArc plans for all the sites.

  17. Derivation of a formula describing the saturation correction of plane-parallel ionization chambers in pulsed fields with arbitrary repetition rate.

    PubMed

    Karsch, Leonhard

    2016-04-21

    Gas-filled ionization chambers are widely used radiation detectors in radiotherapy. A quantitative description and correction of the recombination effects exists for two cases, for continuous radiation exposure and for pulsed radiation fields with short single pulses. This work gives a derivation of a formula for pulsed beams with arbitrary pulse rate for which the prerequisites of the two existing descriptions are not fulfilled. Furthermore, an extension of the validity of the two known cases is investigated. The temporal evolution of idealized charge density distributions within a plane parallel chamber volume is described for pulsed beams of vanishing pulse duration and arbitrary pulse repetition rate. First, the radiation induced release, movement and collection of the charge carriers without recombination are considered. Then, charge recombination is calculated basing on these simplified charge distributions and the time dependent spatial overlap of positive and negative charge carrier distributions. Finally, a formula for the calculation of the saturation correction factor is derived by calculation and simplification of the first two terms of a Taylor expansion for small recombination. The new formula of saturation correction contains the two existing cases, descriptions for exposure by single pulses and continuous irradiation, as limiting cases. Furthermore, it is possible to determine the pulse rate range for which each of the three descriptions is applicable by comparing the dependencies of the new formula with the two existing cases. As long as the time between two pulses is lower than one third of the collection time of the chamber, the formalism for a continuous exposure can be used. The known description for single pulse irradiation is only valid if the repetition rate is less than 1.2 times the inverse collection time. For all other repetition rates in between the new formula should be used. The experimental determination by Jaffe diagrams can be

  18. Dosimetric Verification and Validation of Conformal and IMRT Treatments Fields with an Ionization Chamber 2D-Array

    SciTech Connect

    Evangelina, Figueroa M.; Gabriel, Resendiz G.; Miguel, Perez P.

    2008-08-11

    A three-dimensional treatment planning system requires comparisons of calculated and measured dose distributions. It is necessary to confirm by means of patient specific QA that the dose distributions are correctly calculated, and that the patient data is correctly transferred to and delivered by the treatment machine. We used an analysis software for bi-dimensional dosimetric verification of conformal treatment and IMRT fields using as objective criterion the gamma index. An ionization chamber bi-dimensional array was used for absolute dose measurement in the complete field area.

  19. A method for measuring the electron drift velocity in working gas using a Frisch-grid ionization chamber

    NASA Astrophysics Data System (ADS)

    Bai, Huaiyong; Wang, Zhimin; Zhang, Luyu; Chen, Jinxiang; Zhang, Guohui

    2016-12-01

    A method for measuring the electron drift velocity in working gas is proposed. Based on the cathode and the anode signal waveforms of the Frisch-grid ionization chamber, the electron drift velocity is extracted. With this method, the electron drift velocities in Ar + 10% CH4, Ar + 3.5% CO2 and Kr + 2.7% CO2 gases have been measured and the results are compared with the existing measurements and the simulating results. Using this method, the electron drift velocity can be monitored throughout the experiment of charged particle without bothering the measurement of other parameters, such as the energy and orientation.

  20. Discovery of multiple, ionization-created CS{sub 2} anions and a new mode of operation for drift chambers

    SciTech Connect

    Snowden-Ifft, Daniel P.

    2014-01-15

    This paper focuses on the surprising discovery of multiple species of ionization-created CS{sub 2} anions in gas mixtures containing electronegative CS{sub 2} and O{sub 2}, identified by their slightly different drift velocities. Data are presented to understand the formation mechanism and identity of these new anions. Regardless of the micro-physics, however, this discovery offers a new, trigger-less mode of operation for the drift chambers. A demonstration of trigger-less operation is presented.

  1. Direct determination of k Q for Farmer-type ionization chambers in a clinical scanned carbon ion beam using water calorimetry.

    PubMed

    Osinga-Blättermann, J-M; Brons, S; Greilich, S; Jäkel, O; Krauss, A

    2017-03-21

    Until now, the dosimetry of carbon ions with ionization chambers has not reached the same level of accuracy as that of high-energy photons. This is mainly caused by the approximately threefold larger uncertainty of the k Q factor of ionization chambers, which, due to the lack of experimental data, is still derived by calculations. Measurements of absorbed dose to water, D w, by means of water calorimetry have now been performed in the entrance channel of a scanned 6 cm  ×  6 cm radiation field of 429 MeV/u carbon ions, allowing the direct calibration of ionization chambers and thus the experimental determination of k Q. Within this work, values for k Q have been determined for the Farmer-type ionization chambers FC65-G and TM30013. A detailed investigation of the radiation field enabled the accurate determination of correction factors needed for both calorimetric and ionometric measurements. Finally, a relative standard measurement uncertainty of 0.8% (k  =  1) could be achieved for the experimental k Q values. For both chambers, the experimental k Q factors were found to be about 1% larger than those tabulated in the German DIN 6801-1 protocol, whereas compared to the theoretical values stated in the TRS-398 protocol, the experimental k Q value agrees within 0.4% for the TM30013 chamber but is about 1% lower in the case of the FC65-G chamber.

  2. Direct determination of k Q for Farmer-type ionization chambers in a clinical scanned carbon ion beam using water calorimetry

    NASA Astrophysics Data System (ADS)

    Osinga-Blättermann, J.-M.; Brons, S.; Greilich, S.; Jäkel, O.; Krauss, A.

    2017-03-01

    Until now, the dosimetry of carbon ions with ionization chambers has not reached the same level of accuracy as that of high-energy photons. This is mainly caused by the approximately threefold larger uncertainty of the k Q factor of ionization chambers, which, due to the lack of experimental data, is still derived by calculations. Measurements of absorbed dose to water, D w, by means of water calorimetry have now been performed in the entrance channel of a scanned 6 cm  ×  6 cm radiation field of 429 MeV/u carbon ions, allowing the direct calibration of ionization chambers and thus the experimental determination of k Q. Within this work, values for k Q have been determined for the Farmer-type ionization chambers FC65-G and TM30013. A detailed investigation of the radiation field enabled the accurate determination of correction factors needed for both calorimetric and ionometric measurements. Finally, a relative standard measurement uncertainty of 0.8% (k  =  1) could be achieved for the experimental k Q values. For both chambers, the experimental k Q factors were found to be about 1% larger than those tabulated in the German DIN 6801-1 protocol, whereas compared to the theoretical values stated in the TRS-398 protocol, the experimental k Q value agrees within 0.4% for the TM30013 chamber but is about 1% lower in the case of the FC65-G chamber.

  3. NOTE: The calibration of parallel-plate electron ionization chambers at NPL for use with the IPEM 2003 code of practice: summary data

    NASA Astrophysics Data System (ADS)

    Bass, Graham; Thomas, Russell; Pearce, Julia

    2009-04-01

    The most recent electron dosimetry code of practice for radiotherapy written by the Institute of Physics and Engineering in Medicine was published in 2003 and is based on the NPL electron absorbed dose to water calibration service. NPL has calibrated many Scanditronix type NACP-02 and PTW Roos type 34001 parallel plate ionization chambers in terms of absorbed dose to water, for use with the code of practice. The results of the calibrations of these chamber types summarized here include the absorbed dose to water sensitivity, where the mean calibration factor standard deviations are 5.8% for NACP-02 chambers and 1.1% for PTW Roos chambers. The correction for the polarity effect is shown to be small (less than 0.2% for all beam qualities) but with a discernible beam quality dependence. The correction for recombination is shown to be consistent and reproducible, and an analysis of these results suggests that the plate separation of the NACP-02 chambers is more variable from chamber to chamber than with the PTW Roos chambers. The calibration of these chambers is shown to be repeatable within ±0.2% over 2-3 years. It is also shown that check source measurements can be repeated within ±0.3% over several years. The results justify the use of NACP-02 and PTW 34001 chambers as secondary standards, but also indicate that the PTW 34001 chambers show less variation from chamber to chamber.

  4. The calibration of parallel-plate electron ionization chambers at NPL for use with the IPEM 2003 code of practice: summary data.

    PubMed

    Bass, Graham; Thomas, Russell; Pearce, Julia

    2009-04-21

    The most recent electron dosimetry code of practice for radiotherapy written by the Institute of Physics and Engineering in Medicine was published in 2003 and is based on the NPL electron absorbed dose to water calibration service. NPL has calibrated many Scanditronix type NACP-02 and PTW Roos type 34001 parallel plate ionization chambers in terms of absorbed dose to water, for use with the code of practice. The results of the calibrations of these chamber types summarized here include the absorbed dose to water sensitivity, where the mean calibration factor standard deviations are 5.8% for NACP-02 chambers and 1.1% for PTW Roos chambers. The correction for the polarity effect is shown to be small (less than 0.2% for all beam qualities) but with a discernible beam quality dependence. The correction for recombination is shown to be consistent and reproducible, and an analysis of these results suggests that the plate separation of the NACP-02 chambers is more variable from chamber to chamber than with the PTW Roos chambers. The calibration of these chambers is shown to be repeatable within +/-0.2% over 2-3 years. It is also shown that check source measurements can be repeated within +/-0.3% over several years. The results justify the use of NACP-02 and PTW 34001 chambers as secondary standards, but also indicate that the PTW 34001 chambers show less variation from chamber to chamber.

  5. Analysis of primary aromatic amines in the mainstream waterpipe smoke using liquid chromatography-electrospray ionization tandem mass spectrometry.

    PubMed

    Schubert, Jens; Kappenstein, Oliver; Luch, Andreas; Schulz, Thomas G

    2011-08-19

    In recent years waterpipe smoking has spread worldwide and emerged as global health issue. Yet only little is known on the composition of waterpipe smoke. Here, we present a study on the identification and quantification of primary aromatic amines (PAAs) in this complex environmental matrix. Smoking of the waterpipe was conducted with a smoking machine and particulate matter was collected on glass fiber pads. We developed a fast, simple and specific liquid chromatography-electrospray ionization tandem mass spectrometry (LC-MS/MS) approach to simultaneously detect 31 different PAAs in this matrix. The detection limits comprised a range of 0.45-4.50 ng per smoking session, represented by 2-aminobiphenyl and 3,4,5-trichloroaniline, respectively. Intra- and inter-day precision were determined and proved excellent. We detected 31.3 ± 2.2 ng aniline and 28.0 ± 1.6 ng 4,4'-oxydianiline in the smoke of one waterpipe session. The water in the bowl exerted a small but considerable filter effect on PAAs. The method worked-out showed excellent sensitivity and specificity and is thus highly suited for the determination of PAAs in mainstream waterpipe smoke.

  6. The iQID camera: An ionizing-radiation quantum imaging detector.

    PubMed

    Miller, Brian W; Gregory, Stephanie J; Fuller, Erin S; Barrett, Harrison H; Barber, H Bradford; Furenlid, Lars R

    2014-12-11

    We have developed and tested a novel, ionizing-radiation Quantum Imaging Detector (iQID). This scintillation-based detector was originally developed as a high-resolution gamma-ray imager, called BazookaSPECT, for use in single-photon emission computed tomography (SPECT). Recently, we have investigated the detector's response and imaging potential with other forms of ionizing radiation including alpha, neutron, beta, and fission fragment particles. The confirmed response to this broad range of ionizing radiation has prompted its new title. The principle operation of the iQID camera involves coupling a scintillator to an image intensifier. The scintillation light generated by particle interactions is optically amplified by the intensifier and then re-imaged onto a CCD/CMOS camera sensor. The intensifier provides sufficient optical gain that practically any CCD/CMOS camera can be used to image ionizing radiation. The spatial location and energy of individual particles are estimated on an event-by-event basis in real time using image analysis algorithms on high-performance graphics processing hardware. Distinguishing features of the iQID camera include portability, large active areas, excellent detection efficiency for charged particles, and high spatial resolution (tens of microns). Although modest, iQID has energy resolution that is sufficient to discriminate between particles. Additionally, spatial features of individual events can be used for particle discrimination. An important iQID imaging application that has recently been developed is real-time, single-particle digital autoradiography. We present the latest results and discuss potential applications.

  7. The iQID Camera: An Ionizing-Radiation Quantum Imaging Detector

    DOE PAGES

    Miller, Brian W.; Gregory, Stephanie J.; Fuller, Erin S.; ...

    2014-06-11

    We have developed and tested a novel, ionizing-radiation Quantum Imaging Detector (iQID). This scintillation-based detector was originally developed as a high-resolution gamma-ray imager, called BazookaSPECT, for use in single-photon emission computed tomography (SPECT). Recently, we have investigated the detectors response and imaging potential with other forms of ionizing radiation including alpha, neutron, beta, and fission fragment particles. The detector’s response to a broad range of ionizing radiation has prompted its new title. The principle operation of the iQID camera involves coupling a scintillator to an image intensifier. The scintillation light generated particle interactions is optically amplified by the intensifier andmore » then re-imaged onto a CCD/CMOS camera sensor. The intensifier provides sufficient optical gain that practically any CCD/CMOS camera can be used to image ionizing radiation. Individual particles are identified and their spatial position (to sub-pixel accuracy) and energy are estimated on an event-by-event basis in real time using image analysis algorithms on high-performance graphics processing hardware. Distinguishing features of the iQID camera include portability, large active areas, high sensitivity, and high spatial resolution (tens of microns). Although modest, iQID has energy resolution that is sufficient to discrimate between particles. Additionally, spatial features of individual events can be used for particle discrimination. An important iQID imaging application that has recently been developed is single-particle, real-time digital autoradiography. In conclusion, we present the latest results and discuss potential applications.« less

  8. The iQID Camera: An Ionizing-Radiation Quantum Imaging Detector

    SciTech Connect

    Miller, Brian W.; Gregory, Stephanie J.; Fuller, Erin S.; Barrett, Harrison H.; Barber, Bradford H.; Furenlid, Lars R.

    2014-06-11

    We have developed and tested a novel, ionizing-radiation Quantum Imaging Detector (iQID). This scintillation-based detector was originally developed as a high-resolution gamma-ray imager, called BazookaSPECT, for use in single-photon emission computed tomography (SPECT). Recently, we have investigated the detectors response and imaging potential with other forms of ionizing radiation including alpha, neutron, beta, and fission fragment particles. The detector’s response to a broad range of ionizing radiation has prompted its new title. The principle operation of the iQID camera involves coupling a scintillator to an image intensifier. The scintillation light generated particle interactions is optically amplified by the intensifier and then re-imaged onto a CCD/CMOS camera sensor. The intensifier provides sufficient optical gain that practically any CCD/CMOS camera can be used to image ionizing radiation. Individual particles are identified and their spatial position (to sub-pixel accuracy) and energy are estimated on an event-by-event basis in real time using image analysis algorithms on high-performance graphics processing hardware. Distinguishing features of the iQID camera include portability, large active areas, high sensitivity, and high spatial resolution (tens of microns). Although modest, iQID has energy resolution that is sufficient to discrimate between particles. Additionally, spatial features of individual events can be used for particle discrimination. An important iQID imaging application that has recently been developed is single-particle, real-time digital autoradiography. In conclusion, we present the latest results and discuss potential applications.

  9. The iQID camera: An ionizing-radiation quantum imaging detector

    NASA Astrophysics Data System (ADS)

    Miller, Brian W.; Gregory, Stephanie J.; Fuller, Erin S.; Barrett, Harrison H.; Bradford Barber, H.; Furenlid, Lars R.

    2014-12-01

    We have developed and tested a novel, ionizing-radiation Quantum Imaging Detector (iQID). This scintillation-based detector was originally developed as a high-resolution gamma-ray imager, called BazookaSPECT, for use in single-photon emission computed tomography (SPECT). Recently, we have investigated the detector's response and imaging potential with other forms of ionizing radiation including alpha, neutron, beta, and fission fragment particles. The confirmed response to this broad range of ionizing radiation has prompted its new title. The principle operation of the iQID camera involves coupling a scintillator to an image intensifier. The scintillation light generated by particle interactions is optically amplified by the intensifier and then re-imaged onto a CCD/CMOS camera sensor. The intensifier provides sufficient optical gain that practically any CCD/CMOS camera can be used to image ionizing radiation. The spatial location and energy of individual particles are estimated on an event-by-event basis in real time using image analysis algorithms on high-performance graphics processing hardware. Distinguishing features of the iQID camera include portability, large active areas, excellent detection efficiency for charged particles, and high spatial resolution (tens of microns). Although modest, iQID has energy resolution that is sufficient to discriminate between particles. Additionally, spatial features of individual events can be used for particle discrimination. An important iQID imaging application that has recently been developed is real-time, single-particle digital autoradiography. We present the latest results and discuss potential applications.

  10. The response of smoke detectors to pyrolysis and combustion products from aircraft interior materials

    NASA Technical Reports Server (NTRS)

    Mckee, R. G.; Alvares, N. J.

    1976-01-01

    The following projects were completed as part of the effort to develop and test economically feasible fire-resistant materials for interior furnishings of aircraft as well as detectors of incipient fires in passenger and cargo compartments: (1) determination of the sensitivity of various contemporary gas and smoke detectors to pyrolysis and combustion products from materials commonly used in aircraft interiors and from materials that may be used in the future, (2) assessment of the environmental limitations to detector sensitivity and reliability. The tests were conducted on three groups of materials by exposure to the following three sources of exposure: radiant and Meeker burner flame, heated coil, and radiant source only. The first test series used radiant heat and flame exposures on easily obtainable test materials. Next, four materials were selected from the first group and exposed to an incandescent coil to provide the conditions for smoldering combustion. Finally, radiant heat exposures were used on advanced materials that are not readily available.

  11. DETECTORS AND EXPERIMENTAL METHODS: Software alignment of the BESIII main drift chamber using the Kalman Filter method

    NASA Astrophysics Data System (ADS)

    Wang, Ji-Ke; Mao, Ze-Pu; Bian, Jian-Ming; Cao, Guo-Fu; Cao, Xue-Xiang; Chen, Shen-Jian; Deng, Zi-Yan; Fu, Cheng-Dong; Gao, Yuan-Ning; He, Kang-Lin; He, Miao; Hua, Chun-Fei; Huang, Bin; Huang, Xing-Tao; Ji, Xiao-Bin; Li, Fei; Li, Hai-Bo; Li, Wei-Dong; Liang, Yu-Tie; Liu, Chun-Xiu; Liu, Huai-Min; Liu, Suo; Liu, Ying-Jie; Ma, Qiu-Mei; Ma, Xiang; Mao, Ya-Jun; Mo, Xiao-Hu; Pan, Ming-Hua; Pang, Cai-Ying; Ping, Rong-Gang; Qin, Ya-Hong; Qiu, Jin-Fa; Sun, Sheng-Sen; Sun, Yong-Zhao; Wang, Liang-Liang; Wen, Shuo-Pin; Wu, Ling-Hui; Xie, Yu-Guang; Xu, Min; Yan, Liang; You, Zheng-Yun; Yuan, Chang-Zheng; Yuan, Ye; Zhang, Bing-Yun; Zhang, Chang-Chun; Zhang, Jian-Yong; Zhang, Xue-Yao; Zhang, Yao; Zheng, Yang-Heng; Zhu, Ke-Jun; Zhu, Yong-Sheng; Zhu, Zhi-Li; Zou, Jia-Heng

    2009-03-01

    Software alignment is quite important for a tracking detector to reach its ultimate position accuracy and momentum resolution. We developed a new alignment algorithm for the BESIII Main Drift Chamber using the Kalman Filter method. Two different types of data which are helix tracks and straight tracks are used to test this algorithm, and the results show that the design and implementation is successful.

  12. The coordinate-tracking detector based on the drift chambers for ultrahigh-energy cosmic ray investigations

    NASA Astrophysics Data System (ADS)

    Zadeba, E. A.; Ampilogov, N. V.; Barbashina, N. S.; Bogdanov, A. G.; Borisov, A. A.; Chernov, D. V.; Dushkin, L. I.; Fakhrutdinov, R. M.; Khohlov, S. S.; Kokoulin, R. P.; Kompaniets, K. G.; Kozhin, A. S.; Ovchinnikov, V. V.; Petrukhin, A. A.; Selyakov, V. A.; Shutenko, V. V.; Yashin, I. I.

    2014-08-01

    The project of the tracking detector designed for a joint operation with Cherenkov water detector NEVOD and based on the drift chambers from the neutrino experiment at the IHEP accelerator U-70 is presented. The project is aimed at solving a problem called `muon puzzle' — growing with energy excess of muon content in EAS in comparison with contemporary models of their development, which was registered in various experiments. Joint operation of the coordinate-tracking detector and Cherenkov water calorimeter will allow to measure energy of muon groups and to answer the question about the reasons of the muon flux excess.

  13. General collection efficiency for liquid isooctane and tetramethylsilane used as sensitive media in a parallel-plate ionization chamber.

    PubMed

    Johansson, B; Wickman, G

    1997-01-01

    The general collection efficiency has been measured in liquid isooctane (C8H18) and tetramethylsilane (Si(CH3)4) used as the sensitive media in a parallel-plate ionization chamber, with an electrode distance of 1 mm, intended for photon and electron dosimetry applications. The liquid ionization chamber was irradiated at different dose rates by 140 keV photons from the decay of radioactive 99mTc. The measurements were made at potential differences of 50, 100, 200 and 500 V. Measurements were performed for each liquid and electric field strength, with the decay rate of 99mTc used as the dose-rate reference. The maximum dose rate was about 150 mGy min-1 in each experiment. When the measured general collection efficiency values are compared with the theoretical predictions for collection efficiency in gases, it is found that the latter also describe the general collection efficiency in the two liquids within 1% of the saturation current for collection efficiencies down to 60% when using experimentally determined recombination rate constants and on mobilities characteristic of each of the liquids.

  14. Pulsed discharge helium ionization detector : a new sensitive space detector for gas chromatography ?

    NASA Astrophysics Data System (ADS)

    Szopa, C.; Cabane, M.; Coscia, D.; Coll, P.; Eugenie, J.; Brun, J. F.; Israel, G.

    2003-04-01

    Gas chromatography (GC) is a powerful analytical technique which has been widely used in the exploration of other planetary atmospheres and surfaces. It was part of the scientific payloads devoted to in situ chemical analysis of the soil of Mars, the atmosphere of Venus, and it is currently present in the Huygens probe en route to explore Titan’s atmosphere as well as in the Rosetta lander probe to investigate a cometary nucleus. Obviously, since it was first used in a Viking probe (1976), space GC was improved to fulfil the more and more constraining scientific and space instrumental requirements. More particularly, the separation part, composed of chromatographic columns, was the subject of transformations which contributed to significantly improve the efficiency and the sensitivity of the technique. But on the other hand, space GC detectors remained relatively rudimentary systems, if one except the introduction of mass spectrometry coupled to GC, and they became the limiting factor to a better sensitivity of the whole instrument. In other words, much more sensitive space GC is now required to investigate the composition in organics of hostile environments, as can be the soil of Mars where concentration in organics could be limited to trace levels. That is the reason why, in order to overcome this limitation, we are currently leading a research and development program funded by the space French agency to develop a new type of space chromatographic detector which could meet the objectives of the future space GC : the pulsed discharge helium ionisation detector (PDHID). The aim of this paper is to demonstrate the compatibility of the analytical performances of PDHID with space exploration requirements, by presenting the results of a series of experiments led with a commercial version of PDHID. These results made us starting a program of miniaturisation of this type of detector to build a version compatible with space instrumentation requirements in terms of mass

  15. Influence of solvothermal synthesis conditions in BiSI nanostructures for application in ionizing radiation detectors

    NASA Astrophysics Data System (ADS)

    Aguiar, I.; Mombrú, M.; Pérez Barthaburu, M.; Bentos Pereira, H.; Fornaro, L.

    2016-02-01

    BiSI belongs to the A V B VI C VII chalcohalides group of compounds. These compounds show several interesting properties such as ferroelectricity, piezoelectricity along the c axis, and photoconductivity. Moreover, BiSI is a potential semiconductor material for room-temperature gamma and x-ray detection, given its band gap of 1.57 eV and its high density, 6.41 g cm-3. In this work we present BiSI nanostructures synthesized by the solvothermal method with the intention of using them for ionizing radiation detection. The solvent was varied to study its influence in morphology, particle size and size distribution. Three different conditions were tested, using either water, monoethylene glycol and a mixture of both solvents. Nanostructures were characterized by XRD to determine the phase obtained and reaction completeness; TEM was used to observe nanostructures morphology, size, size distribution and crystallinity; and finally FT-IR diffuse reflectance was used to study monoethylene glycol presence in the samples. Nanorods in the range of 100-200 nm width were obtained in all samples, but round nanoparticles of around 10 nm in diameter were also detected in samples synthesized only with monoethylene glycol. Samples synthesized in monoethylene glycol were used to fabricate pellets to construct detectors. The detectors responded to ionizing radiation and a resistivity in the order of 1013 Ω cm was estimated. This work proposes, to our knowledge, the first study of BiSI for its application in ionizing radiation detection.

  16. Numeric estimation of the possibilities of ionizing detectors for monitoring SR beam space parameters

    NASA Astrophysics Data System (ADS)

    Artemiev, A. N.; Artemiev, N. A.; Ioudin, L. I.; Mikhailov, V. G.; Moryakov, V. P.; Odintsov, D. G.; Rezvov, V. A.; Cerenius, Y.; Svensson, A.

    2000-06-01

    An ionizing detector for on-line registration and representation of the geometric SR beam parameters was developed in RRC KI. The detector analyses the products of the residual gas ionization, which was done by the investigated beam. Special electrostatic optics and open image converter tube (ICT) form optical image of the real beam on the screen of ICT. The detector was checked on SR beams of the next storage rings: DCI (LURE, Orsey, France), KSRS (RRC KI, Moscow, Russia) and MAX-2 (MAX-lab, Lund, Sweden). The codes for TV image processing give a possibility for numeric estimation of the beam size, the width of its horizontal and vertical profiles and position of the beam gravity. Statistic processing of the beam gravity center using big amount of TV frames gives uncertainty in the beam position of about 2 μm while the width of the beam is about 2 mm. Summation of big amount of TV frames was used. This method significantly increases signal-to-noise ratio.

  17. SU-D-19A-01: Can Farmer-Type Ionization Chambers Be Used to Improve the Accuracy of Low-Energy Electron Beam Reference Dosimetry?

    SciTech Connect

    Muir, B R; McEwen, M R

    2014-06-01

    Purpose: To investigate the use of cylindrical Farmer-type ionization chambers to improve the accuracy of low-energy electron beam calibration. Historically, these chamber types have not been used in beams with incident energies less than 10 MeV (R{sub 5} {sub 0} < 4.3 cm) because early investigations suggested large (up to 5 %) fluence perturbation factors in these beams, implying that a significant component of uncertainty would be introduced if used for calibration. More recently, the assumptions used to determine perturbation corrections for cylindrical chambers have been questioned. Methods: Measurements are made with cylindrical chambers in Elekta Precise 4, 8 and 18 MeV electron beams. Several chamber types are investigated that employ graphite walls and aluminum electrodes with very similar specifications (NE2571, NE2505/3, FC65-G). Depth-ionization scans are measured in water in the 8 and 18 MeV beams. To reduce uncertainty from chamber positioning, measurements in the 4 MeV beam are made at the reference depth in Virtual Water™. The variability of perturbation factors is quantified by comparing normalized response of various chambers. Results: Normalized ion chamber response varies by less than 0.7 % for similar chambers at average electron energies corresponding to that at the reference depth from 4 or 6 MeV beams. Similarly, normalized measurements made with similar chambers at the reference depth in the 4 MeV beam vary by less than 0.4 %. Absorbed dose calibration coefficients derived from these results are stable within 0.1 % on average over a period of 6 years. Conclusion: These results indicate that the uncertainty associated with differences in fluence perturbations for cylindrical chambers with similar specifications is only 0.2 %. The excellent long-term stability of these chambers in both photon and electron beams suggests that these chambers might offer the best performance for all reference dosimetry applications.

  18. The iQID camera: An ionizing-radiation quantum imaging detector

    PubMed Central

    Miller, Brian W.; Gregory, Stephanie J.; Fuller, Erin S.; Barrett, Harrison H.; Barber, H. Bradford; Furenlid, Lars R.

    2015-01-01

    We have developed and tested a novel, ionizing-radiation Quantum Imaging Detector (iQID). This scintillation-based detector was originally developed as a high-resolution gamma-ray imager, called BazookaSPECT, for use in single-photon emission computed tomography (SPECT). Recently, we have investigated the detector’s response and imaging potential with other forms of ionizing radiation including alpha, neutron, beta, and fission fragment particles. The confirmed response to this broad range of ionizing radiation has prompted its new title. The principle operation of the iQID camera involves coupling a scintillator to an image intensifier. The scintillation light generated by particle interactions is optically amplified by the intensifier and then re-imaged onto a CCD/CMOS camera sensor. The intensifier provides sufficient optical gain that practically any CCD/CMOS camera can be used to image ionizing radiation. The spatial location and energy of individual particles are estimated on an event-by-event basis in real time using image analysis algorithms on high-performance graphics processing hardware. Distinguishing features of the iQID camera include portability, large active areas, excellent detection efficiency for charged particles, and high spatial resolution (tens of microns). Although modest, iQID has energy resolution that is sufficient to discriminate between particles. Additionally, spatial features of individual events can be used for particle discrimination. An important iQID imaging application that has recently been developed is real-time, single-particle digital autoradiography. We present the latest results and discuss potential applications. PMID:26166921

  19. Pulse-Shape Analysis of Ionization Signals in Cryogenic Ge Detectors for Dark Matter

    NASA Astrophysics Data System (ADS)

    Foerster, N.; Broniatowski, A.; Eitel, K.; Marnieros, S.; Paul, B.; Piro, M.-C.; Siebenborn, B.

    2016-08-01

    The detectors of the direct dark matter search experiment EDELWEISS consist of high-purity germanium crystals operated at cryogenic temperatures (mathrm {{<}20 mK}) and low electric fields (mathrm {{<}1 V/cm}). The surface discrimination is based on the simultaneous measurement of the charge amplitudes on different sets of electrodes. As the rise time of a charge signal strongly depends on the location of an interaction in the crystal, a time-resolved measurement can also be used to identify surface interactions. This contribution presents the results of a study of the discrimination power of the rise time parameter from a hot carrier transport simulation in combination with time-resolved measurements using an EDELWEISS-type detector in a test cryostat at ground level. We show the setup for the time-resolved ionization signal read-out in the EDELWEISS-III experiment and first results from data taking in the underground laboratory of Modane.

  20. Study of timing response and charge spectra of glass based Resistive Plate Chamber detectors for INO-ICAL experiment

    NASA Astrophysics Data System (ADS)

    Gaur, A.; Kumar, A.; Naimuddin, Md.

    2017-03-01

    Resistive Plate chambers (RPCs) are robust and affordable gaseous detectors that combine low cost with excellent timing, good spatial resolution and fast response to the incoming particles. The India Based Neutrino Observatory is an approved project aimed at building a magnetised Iron Calorimeter (ICAL) detector to study Neutrino physics and related issues. The ICAL experiment will utilize about 29000 RPC's as active detector elements, sandwiched between alternate plates of thick iron. The RPC detectors will be used to detect muons produced from the atmospheric neutrinos interaction with an iron target. The spatial information of the muons will be extracted from the two dimensional readout and the hit position in the respective layers. The up going and down going directionality will be obtained using the time stamp of hits in the active detectors. The charge induced by the particle and its behaviour with respect to the applied voltage play a significant role in designing the readout electronics for the detector. In this paper, we present the timing and charge measurement of single gap glass based RPC detectors. We will also report about studies on the dependence of the timing and charge response of these RPC detectors as a function of the gas composition.

  1. Atmospheric measurements by Medipix-2 and Timepix Ionizing Radiation Imaging Detectors on BEXUS stratospheric balloon campaigns

    NASA Astrophysics Data System (ADS)

    Urbar, Jaroslav; Scheirich, Jan; Jakubek, Jan

    2010-05-01

    Results of the first two experiments using semiconductor pixel detectors of the Medipix family for cosmic ray imaging in the stratospheric environment are presented. The original detecting device was based on the hybrid pixel detectors of Medipix-2 and Timepix developed at CERN with USB interface developed at Institute of Experimental and Applied Physics of Czech Technical University in Prague. The detectors were used in tracking mode allowing them to operate as an "active nuclear emulsion". The actual flight time of BEXUS7 with Medipix-2 on 8th October 2008 was over 4 hours, with 2 hours at stable floating altitude of 26km. BEXUS9 measurements of similar duration by Timepix, Medipix-2 and ST-6 Geiger telescope instruments took place in arctic atmosphere below 24km altitude on 11th October 2009. This balloon platform is quite ideal for such in-situ measurements. Not only because of the high altitudes reached, but also due to its slow ascent velocity for statistically relevant sampling of the ambient environment for improving cosmic ray induced ionisation rate model inputs. The flight opportunity for BEXUS student projects was provided by Education department of the European Space Agency (ESA) and Eurolaunch - Collaboration of Swedish National Space Board (SNSB) and German Space Agency (DLR). The scientific goal was to check energetic particle type altitudinal dependencies, also testing proper detector calibration by detecting fluxes of ionizing radiation, while evaluating instrumentation endurance and performance.

  2. A high rate proportional chamber

    SciTech Connect

    Henderson, R.; Fraszer, W.; Openshaw, R.; Sheffer, G.; Salomon, M.; Dew, S.; Marans, J.; Wilson, P.

    1987-02-01

    Gas mixtures with high specific ionization allow the use of small interelectrode distances while still maintaining full efficiency. With the short electron drift distances the timing resolution is also improved. The authors have built and operated two 25 cm/sup 2/ chambers with small interelectrode distances. Also single wire detector cells have been built to test gas mixture lifetimes. Various admixtures of CF/sub 4/, DME, Isobutane, Ethane and Argon have been tested. Possible applications of such chambers are as beam profile monitors, position tagging of rare events and front end chambers in spectrometers.

  3. Simulation of the Mg(Ar) ionization chamber currents by different Monte Carlo codes in benchmark gamma fields

    NASA Astrophysics Data System (ADS)

    Lin, Yi-Chun; Liu, Yuan-Hao; Nievaart, Sander; Chen, Yen-Fu; Wu, Shu-Wei; Chou, Wen-Tsae; Jiang, Shiang-Huei

    2011-10-01

    High energy photon (over 10 MeV) and neutron beams adopted in radiobiology and radiotherapy always produce mixed neutron/gamma-ray fields. The Mg(Ar) ionization chambers are commonly applied to determine the gamma-ray dose because of its neutron insensitive characteristic. Nowadays, many perturbation corrections for accurate dose estimation and lots of treatment planning systems are based on Monte Carlo technique. The Monte Carlo codes EGSnrc, FLUKA, GEANT4, MCNP5, and MCNPX were used to evaluate energy dependent response functions of the Exradin M2 Mg(Ar) ionization chamber to a parallel photon beam with mono-energies from 20 keV to 20 MeV. For the sake of validation, measurements were carefully performed in well-defined (a) primary M-100 X-ray calibration field, (b) primary 60Co calibration beam, (c) 6-MV, and (d) 10-MV therapeutic beams in hospital. At energy region below 100 keV, MCNP5 and MCNPX both had lower responses than other codes. For energies above 1 MeV, the MCNP ITS-mode greatly resembled other three codes and the differences were within 5%. Comparing to the measured currents, MCNP5 and MCNPX using ITS-mode had perfect agreement with the 60Co, and 10-MV beams. But at X-ray energy region, the derivations reached 17%. This work shows us a better insight into the performance of different Monte Carlo codes in photon-electron transport calculation. Regarding the application of the mixed field dosimetry like BNCT, MCNP with ITS-mode is recognized as the most suitable tool by this work.

  4. Ultra-thin plasma radiation detector

    DOEpatents

    Friedman, Peter S.

    2017-01-24

    A position-sensitive ionizing-radiation counting detector includes a radiation detector gas chamber having at least one ultra-thin chamber window and an ultra-thin first substrate contained within the gas chamber. The detector further includes a second substrate generally parallel to and coupled to the first substrate and defining a gas gap between the first substrate and the second substrate. The detector further includes a discharge gas between the substrates and contained within the gas chamber, where the discharge gas is free to circulate within the gas chamber and between the first and second substrates at a given gas pressure. The detector further includes a first electrode coupled to one of the substrates and a second electrode electrically coupled to the first electrode. The detector further includes a first discharge event detector coupled to at least one of the electrodes for detecting a gas discharge counting event in the electrode.

  5. Smoking

    MedlinePlus

    ... harms nearly every organ of the body. Cigarette smoking causes 87 percent of lung cancer deaths. It is also responsible for many other cancers ... or having a baby die from sudden infant death syndrome (SIDS). Your smoke is ... are battery-operated smoking devices. Not much is known about the health ...

  6. Fast-Neutron Spectrometry Using a 3He Ionization Chamber and Digital Pulse Shape Analysis

    SciTech Connect

    D. L. Chichester; J. T. Johnson; E. H. Seabury

    2010-05-01

    Digital pulse shape analysis (dPSA) has been used with a Cuttler-Shalev type 3He proportional counter to measure the fast neutron spectra of bare 252Cf and 241AmBe neutron sources. Measurements have also been made to determine the attenuated fast neutron spectra of 252Cf shielded by several materials including water, graphite, liquid nitrogen, magnesium, and tungsten. Rise-time dPSA has been employed using the common rise-time approach for analyzing n +3He ? 1H + 3H ionization events and a new approach has been developed to improve the fidelity of these measurements. Simulations have been performed for the different experimental arrangements and are compared, demonstrating general agreement between the dPSA processed fast neutron spectra and predictions.

  7. On the effect of temperature and the width of the turbulent combustion zone on the ionization detector readings

    NASA Astrophysics Data System (ADS)

    Shaikin, A. P.; Galiev, I. R.

    2016-08-01

    We have considered the functional dependence of the ionization detector readings (ion current) on the composition of the fuel-air mixture, adiabatic temperature, and the turbulent combustion zone width. Experiments on the engine show that, for an air excess factor of 0.75-1.15, the coincidence of the calculated and experimental data exceeds 90%. Our results can be used to predict and monitor the adiabatic temperature of the flame and the width of the turbulent combustion zone in the combustion changer of the internal combustion engine using the ionization detector.

  8. Ionisation chamber containing boron as a neutron detector in medical accelerator fields.

    PubMed

    Zielczynski, M; Gryzinski, M A; Golnik, N; Tulik, P

    2007-01-01

    A combination of the recombination principle of H(10) measurements with the use of the ionisation chambers containing boron has been presented, in order to increase the relative sensitivity of the chamber to neutrons by a factor close to the radiation quality factor of photoneutrons. Three types of the chambers were investigated. Two of them were filled with BF(3) and the third one contained electrodes covered with B(4)C. All the chambers were placed in paraffin moderators. The response of the chambers was investigated, depending on gas pressure and polarising voltage. The results showed that it was possible to obtain nearly the same response of the chamber to H(10) for photons and neutrons in a restricted energy range; however, further investigations are needed to make an optimum design.

  9. A micro-gap, air-filled ionisation chamber as a detector for criticality accident dosimetry.

    PubMed

    Murawski, Ł; Zielczyński, M; Golnik, N; Gryziński, M A

    2014-10-01

    A micro-gap air-filled ionisation chamber was designed for criticality dosimetry. The special feature of the chamber is its very small gap between electrodes of only 0.3 mm. This prevents ion recombination at high dose rates and minimises the influence of gas on secondary particles spectrum. The electrodes are made of polypropylene because of higher content of hydrogen in this material, when compared with soft tissue. The difference between neutron and gamma sensitivity in such chamber becomes practically negligible. The chamber's envelope contains two specially connected capacitors, one for polarising the electrodes and the other for collecting the ionisation charge.

  10. Laser Induced Fluorescence Spectroscopy of Neutral and Ionized Polycyclic Aromatic Hydrocarbons in the Cosmic Simulation Chamber

    NASA Technical Reports Server (NTRS)

    Bejaoui, Salma; Salama, Farid; Contreras, Cesar; Sciamma O'Brien, Ella; Foing, Bernard; Pascale, Ehrenfreund

    2015-01-01

    Polycyclic aromatic hydrocarbon (PAH) molecules are considered the best carriers to account for the ubiquitous infrared emission bands. PAHs have also been proposed as candidates to explain the diffuse interstellar bands (DIBs), a series of absorption features seen on the interstellar extinction curve and are plausible carriers for the extended red emission (ERE), a photoluminescent process associated with a wide variety of interstellar environments. Extensive efforts have been devoted over the past two decades to characterize the physical and chemical properties of PAH molecules and ions in space. Absorption spectra of PAH molecules and ions trapped in solid matrices have been compared to the DIBs. Absorption spectra of several cold, isolated gas-phase PAHs have also been measured under experimental conditions that mimic the interstellar conditions. The purpose of this study is to provide a new dimension to the existing spectroscopic database of neutral and single ionized PAHs that is largely based on absorption spectra by adding emission spectroscopy data. The measurements are based on the laser induced fluorescence (LIF) technique and are performed with the Pulsed Discharge Nozzle (PDN) of the COSmIC laboratory facility at NASA Ames laboratory. The PDN generates a plasma in a free supersonic jet expansion to simulate the physical and the chemical conditions in interstellar environments. We focus, here, on the fluorescence spectra of large neutral PAHs and their cations where there is a lack of fluorescence spectroscopy data. The astronomical implications of the data (e.g., ERE) are examined.

  11. A pyroloysis technique for determining microamounts of hydrogen in lunar soil using the helium ionization detector

    NASA Technical Reports Server (NTRS)

    Bustin, R.

    1983-01-01

    A method has been developed which will determine hydrogen in sub-milligram samples of lunar soil. It consists of heating the sample in a pyroprobe followed by the gas chromatographic determination of hydrogen using the helium ionization detector. Using a 7 foot, 1/8 OD stainless steel column packed with Carbosieve S, 120/140 mesh, hydrogen was well-separated from the other gases released from lunar soil. Standards of hydrogen in helium were used for calibration. The limit to detection under the conditions used was about 2 ng. The method was linear from 2 ng to 270 ng. The method was checked using some actual lunar samples. Results were typical of those obtained for lunar soils using other methods.

  12. SU-E-T-225: Correction Matrix for PinPoint Ionization Chamber for Dosimetric Measurements in the Newly Released Incise™ Multileaf Collimator Shaped Small Field for CyberKnife M6™ Machine

    SciTech Connect

    Zhang, Y; Li, T; Heron, D; Huq, M

    2015-06-15

    Purpose: For small field dosimetry, such as measurements of output factors for cones or MLC-shaped irregular small fields, ion chambers often Result in an underestimation of the dose, due to both the volume averaging effect and the lack of lateral charged particle equilibrium. This work presents a mathematical model for correction matrix for a PTW PinPoint ionization chamber for dosimetric measurements made in the newly released Incise™ Multileaf collimator fields of the CyberKnife M6™ machine. Methods: A correction matrix for a PTW 0.015cc PinPoint ionization chamber was developed by modeling its 3D dose response in twelve cone-shaped circular fields created using the 5mm, 7.5mm, 10mm, 12.5mm, 15mm, 20mm, 25mm, 30mm, 35mm, 40mm, 50mm, 60mm cones in a CyberKnife M6™ machine. For each field size, hundreds of readings were recorded for every 2mm chamber shift in the horizontal plane. The contribution of each dose pixel to a measurement point depended on the radial distance and the angle to the chamber axis. These readings were then compared with the theoretical dose as obtained with Monte Carlo calculation. A penalized least-square optimization algorithm was developed to generate the correction matrix. After the parameter fitting, the mathematical model was validated for MLC-shaped irregular fields. Results: The optimization algorithm used for parameter fitting was stable and the resulted response factors were smooth in spatial domain. After correction with the mathematical model, the chamber reading matched with the calculation for all the tested fields to within 2%. Conclusion: A novel mathematical model has been developed for PinPoint chamber for dosimetric measurements in small MLC-shaped irregular fields. The correction matrix is dependent on detector, treatment unit and the geometry of setup. The model can be applied to non-standard composite fields and provides an access to IMRT point dose validation.

  13. NOTE: Radiological thickness measurement using a liquid ionization chamber electronic portal imaging device

    NASA Astrophysics Data System (ADS)

    Evans, Philip M.; Donovan, Ellen M.; Partridge, Mike; Bidmead, A. Margaret; Garton, Andrew; Mubata, Cephas

    1999-06-01

    We present a method of calibrating the Portal Vision electronic portal imaging device to obtain radiological thickness maps for compensator design. In this method, coefficients are derived to describe the relationship between intensity and thickness for a set of water-equivalent blocks. The effects of four parameters were studied: (a) The dose response of the system was measured and found to be describable by a square-root function. (b) The calibration data and images were taken with a wedge in situ. The effects of using different wedges and different wedge orientations were investigated. The intrinsic accuracy of the accelerator/imager system was found to be 1.9 mm, for both 15° and 30° wedges. Changing the wedge orientation between calibration and imaging and rotating the calibration coefficients accordingly led to an error of 3.5 mm. (c) The variation in detector response with gantry angle was measured and corrected. The residual error in this process was 2.4 mm. (d) The use of a model to correct the effects of imaging with different field sizes was investigated and found to yield a residual error of 2.9 mm. The overall error in image calibrations was approx 4 mm or 2% in dose. This is considered to be sufficiently small for the intended use of designing compensators for tangential breast irradiation.

  14. Laser Induced Fluorescence Spectroscopy of Neutral and Ionized Polycyclic Aromatic Hydrocarbons in a Cosmic Simulation Chamber

    NASA Astrophysics Data System (ADS)

    Bejaoui, Salma; Salama, Farid

    2015-08-01

    Polycyclic aromatic hydrocarbon (PAH) molecules are considered the best carriers to account for the ubiquitous infrared emission bands. PAHs have also been proposed as candidates to explain the diffuse interstellar bands (DIBs), a series of absorption features seen on the interstellar extinction curve and are plausible carriers for the extended red emission (ERE), a photoluminescent process associated with a wide variety of interstellar environments. Extensive efforts have been devoted over the past two decades to characterize the physical and chemical properties of PAH molecules and ions in space. Absorption spectra of PAH molecules and ions trapped in solid matrices have been compared to the DIBs [1, 2]. Absorption spectra of several cold, isolated gas-phase PAHs have also been measured under experimental conditions that mimic the interstellar conditions [see 3 for a review]. The purpose of this study is to provide a new dimension to the existing spectroscopic database of neutral and single ionized PAHs that is largely based on absorption spectra by adding emission spectroscopy data. The measurements are based on the laser-induced fluorescence (LIF) technique [4] and are performed with the Pulsed Discharge Nozzle (PDN) of the COSmIC laboratory facility at NASA Ames laboratory. The PDN generates plasma in a free supersonic jet expansion to simulate the physical and the chemical conditions in interstellar environments. We focus, here, on the fluorescence spectra of large neutral PAHs and their cations where there is a lack of fluorescence spectroscopy data. The astronomical implications of the data (e.g., ERE) are examinedReferences[1] F. Salama, E. Bakes, L.J. Allamandola, A.G.G.M. Tielens, Astrophys. J., 458 (1996) p.621[2] F. Salama, The ISO Revolution, EDP Sciences, Les Ulis, France (1999) p.65[3] Salama F., In Organic Matter in Space, IAU Symposium 251, Kwok & Sandford Eds.Cambridge University Press,4, S251,(2008), p. 357 (2008) and references therein.[4

  15. On the interpretation of current-voltage curves in ionization chambers using the exact solution of the Thomson problem

    NASA Astrophysics Data System (ADS)

    Ridenti, M. A.; Pascholati, P. R.; Gonçalves, J. A. C.; Bueno, C. C.

    2015-09-01

    The I - ΔV characteristic curve of a well type ionization chamber irradiated with 192Ir sources (0.75 Ci-120 Ci) was fitted using the exact solution of the Thomson problem. The recombination coefficient and saturation current were estimated using this new approach. The saturation current was compared with the results of the conventional method based on Boag-Wilson formula. It was verified that differences larger than 1% between both methods only occurred at activities higher than 55 Ci. We concluded that this new approach is recommended for a more accurate estimate of the saturation current when it is not possible to measure currents satisfying the condition I /Isat > 0.95. From the calibration curve the average value of pairs of carriers created per unit volume was estimated to be equal to η = 8.1 ×10-3cm-3s-1 Bq-1 and from that value it was estimated that ~ 17 pairs were created on average per second for each decay of the source.

  16. Measurement of absorbed dose-to-water for an HDR {sup 192}Ir source with ionization chambers in a sandwich setup

    SciTech Connect

    Araki, Fujio; Kouno, Tomohiro; Ohno, Takeshi; Kakei, Kiyotaka; Yoshiyama, Fumiaki; Kawamura, Shinji

    2013-09-15

    Purpose: In this study, a dedicated device for ion chamber measurements of absorbed dose-to-water for a Nucletron microSelectron-v2 HDR {sup 192}Ir brachytherapy source is presented. The device uses two ionization chambers in a so-called sandwich assembly. Using this setup and by taking the average reading of the two chambers, any dose error due to difficulties in absolute positioning (centering) of the source in between the chambers is cancelled to first order. The method's accuracy was examined by comparing measurements with absorbed dose-to-water determination based on the AAPM TG-43 protocol.Methods: The optimal source-to-chamber distance (SCD) for {sup 192}Ir dosimetry was determined from ion chamber measurements in a water phantom. The {sup 192}Ir source was sandwiched between two Exradin A1SL chambers (0.057 cm{sup 3}) at the optimal SCD separation. The measured ionization was converted to the absorbed dose-to-water using a {sup 60}Co calibration factor and a Monte Carlo-calculated beam quality conversion factor, k{sub Q}, for {sup 60}Co to {sup 192}Ir. An uncertainty estimate of the proposed method was determined based on reproducibility of measurements at different institutions for the same type of source.Results: The optimal distance for the A1SL chamber measurements was determined to be 5 cm from the {sup 192}Ir source center, considering the depth dependency of k{sub Q} for {sup 60}Co to {sup 192}Ir and the chamber positioning. The absorbed dose to water measured at (5 cm, 90°) on the transverse axis was 1.3% lower than TG-43 values and its reproducibility and overall uncertainty were 0.8% and 1.7%, respectively. The measurement doses at anisotropic points agreed within 1.5% with TG-43 values.Conclusions: The ion chamber measurement of absorbed dose-to-water with a sandwich method for the {sup 192}Ir source provides a more accurate, direct, and reference dose compared to the dose-to-water determination based on air-kerma strength in the TG-43 protocol

  17. SU-E-T-561: Development of Depth Dose Measurement Technique Using the Multilayer Ionization Chamber for Spot Scanning Method

    SciTech Connect

    Takayanagi, T; Fujitaka, S; Umezawa, M; Ito, Y; Nakashima, C; Matsuda, K

    2014-06-01

    Purpose: To develop a measurement technique which suppresses the difference between profiles obtained with a multilayer ionization chamber (MLIC) and with a water phantom. Methods: The developed technique multiplies the raw MLIC data by a correction factor that depends on the initial beam range and water equivalent depth. The correction factor is derived based on a Bragg curve calculation formula considering range straggling and fluence loss caused by nuclear reactions. Furthermore, the correction factor is adjusted based on several integrated depth doses measured with a water phantom and the MLIC. The measured depth dose profiles along the central axis of the proton field with a nominal field size of 10 by 10 cm were compared between the MLIC using the new technique and the water phantom. The spread out Bragg peak was 20 cm for fields with a range of 30.6 cm and 6.9 cm. Raw MLIC data were obtained with each energy layer, and integrated after multiplying by the correction factor. The measurements were performed by a spot scanning nozzle at Nagoya Proton Therapy Center, Japan. Results: The profile measured with the MLIC using the new technique is consistent with that of the water phantom. Moreover, 97% of the points passed the 1% dose /1mm distance agreement criterion of the gamma index. Conclusion: We have demonstrated that the new technique suppresses the difference between profiles obtained with the MLIC and with the water phantom. It was concluded that this technique is useful for depth dose measurement in proton spot scanning method.

  18. Data-driven exploration of the ionization-phonon partitioning in scintillating radiation detector materials

    SciTech Connect

    Ferris, Kim F.; Webb-Robertson, Bobbie-Jo M.; Jordan, David V.; Jones, Dumont M.

    2008-06-01

    An information-based approach to scintillating materials development has been applied to ranking the alkali halide and alkali earth halide series in terms of their energy conversion efficiency. The efficiency of scintillating radiation detection materials can be viewed as the product of a consecutive series of electronic processes (energy conversion, transfer, and luminescence) as outlined by Lempicki and others. Relevant data are relatively sparse, but sufficient for the development of forward mapping of materials properties through materials signatures. These mappings have been used to explore the limits of the K ratio in the Lempicki model with chemical composition, and examine its relationship with another common design objective, density. The alkali halides and alkali earth halide compounds separate themselves into distinct behavior classes favoring heavier cations and anions for improved K ratio. While the coupling of ionization is strongly related to the optical phonon modes, both dielectric and band gap contributions cannot be ignored. When applied within a candidate screen, the resulting model for K imposes design rules—simple structural restrictions—on scintillating radiation detector materials.

  19. Improved gas chromatography-flame ionization detector analytical method for the analysis of epoxy fatty acids.

    PubMed

    Mubiru, Edward; Shrestha, Kshitij; Papastergiadis, Antonios; De Meulenaer, Bruno

    2013-11-29

    In this study an improved method for analysis of epoxy fatty acids is reported. Data obtained from analysis of polar fatty acids has previously been presented, but due to the high number of compounds that co-elute in the polar fraction, the resultant chromatograms are complex which may lead to compromising the accuracy of the data. A three steps separation of fatty acid methyl esters (FAMEs) by solid-phase extraction (SPE) on a silica gel column to remove hydroxy fatty acid interferences was proposed. This approach is opposed to a two step separation procedure that has been often used to prevent analytical interferences caused by non-altered fatty acids. A gas chromatograph with a flame ionization detector (GC-FID) equipped with a polar CP-Sil 88™ column was used. Quantification was based on the use of methyl nonadecanoate (C19:0), as an internal standard. Individual mono epoxy fatty acids were well separated without co-eluting compounds. The optimized method was finally applied to screen epoxy fatty acids in 37 fresh oil samples. Results obtained for the total epoxy fatty acids were in the range 0.03-2mgg(-1) of oil with repeatability coefficient of variation (CV) ranging from 2.8 to 9.9% for duplicate analysis showing that the results obtained are repeatable.

  20. SU-D-213-04: Accounting for Volume Averaging and Material Composition Effects in An Ionization Chamber Array for Patient Specific QA

    SciTech Connect

    Fugal, M; McDonald, D; Jacqmin, D; Koch, N; Ellis, A; Peng, J; Ashenafi, M; Vanek, K

    2015-06-15

    Purpose: This study explores novel methods to address two significant challenges affecting measurement of patient-specific quality assurance (QA) with IBA’s Matrixx Evolution™ ionization chamber array. First, dose calculation algorithms often struggle to accurately determine dose to the chamber array due to CT artifact and algorithm limitations. Second, finite chamber size and volume averaging effects cause additional deviation from the calculated dose. Methods: QA measurements were taken with the Matrixx positioned on the treatment table in a solid-water Multi-Cube™ phantom. To reduce the effect of CT artifact, the Matrixx CT image set was masked with appropriate materials and densities. Individual ionization chambers were masked as air, while the high-z electronic backplane and remaining solid-water material were masked as aluminum and water, respectively. Dose calculation was done using Varian’s Acuros XB™ (V11) algorithm, which is capable of predicting dose more accurately in non-biologic materials due to its consideration of each material’s atomic properties. Finally, the exported TPS dose was processed using an in-house algorithm (MATLAB) to assign the volume averaged TPS dose to each element of a corresponding 2-D matrix. This matrix was used for comparison with the measured dose. Square fields at regularly-spaced gantry angles, as well as selected patient plans were analyzed. Results: Analyzed plans showed improved agreement, with the average gamma passing rate increasing from 94 to 98%. Correction factors necessary for chamber angular dependence were reduced by 67% compared to factors measured previously, indicating that previously measured factors corrected for dose calculation errors in addition to true chamber angular dependence. Conclusion: By comparing volume averaged dose, calculated with a capable dose engine, on a phantom masked with correct materials and densities, QA results obtained with the Matrixx Evolution™ can be significantly

  1. D-Zero General Support: Description of the D-Zero High Sensitivity Smoke Detector System (HSSD)

    SciTech Connect

    Hance, Rick; /Fermilab

    2001-01-25

    There are nine separate zones of high sensitivity smoke detection (HSSD) at Dzero. These zones, their coverage, their condition thresholds, and their response to conditions are described in the table in the 'Zones & Responses' section of this document. This is an 'air-sampling' smoke detection system. Each zone is formed of a network of pipes which contain strategically sized and located air sampling holes. Each zone has a fanbox/detector head that continuously draws air into the holes and through the pipes to the detector head where it is evaluated for combustion products. The fanbox/detector heads are located on the Dzero detector platform and in the trusses. Each fanbox/detector head has a remotely located display control card (DCC) for monitoring the level of obscuration of the air sample. This remote equipment is located on the roof of the moving counting house. The DCC for each zone is networked to a single intelligent interface module (11M), also on the roof of the MCH, that collects the information from all nine zones and delivers it to the Dzero control room via an isolated RS-485 link to an operators console. This console is replicated via a 'remote video, keyboard, mouse system' in the DAB 4th floor entry lobby for use by fire response personnel. Internal thresholds that have been set in the DCC via the operators console provides alarm level I (lowest), alarm level 2, and alarm level 3 (highest) for each zone. These alarms are then used as inputs to the Fermilab FIRUS system, and as interlocks to equipment. This system is nearly 2000 times more sensitive than conventional smoke detectors. It is sensitive enough to detect invisible products of combustion such as the outgassing of electrical components, overheating PVC wire insulation, or small changes in the ambient level of obscuration caused in the incipient stage of a fire.

  2. Analytical solutions of minimum ionization particle induced current shapes of silicon detectors and simulation of charge collection properties

    SciTech Connect

    Eremin, V.; Chen, W.; Li, Z.

    1993-11-01

    A new analytical, one dimensional method to obtain the induced current shapes and simulation of chasrge shapes for p{sup +} {minus}n{minus}n{sup +} silicon detectors in the case of minimum ionization particle has been developed here. jExact solutions have been found for both electron and hole current shapes. Simulations of induced charge shapes of detectors have also been given. The results of this work are consistent with the earlier work where a semi-analytical method had been used.

  3. SU-E-T-645: Qualification of a 2D Ionization Chamber Array for Beam Steering and Profile Measurement

    SciTech Connect

    Gao, S; Balter, P; Rose, M; Simon, W

    2015-06-15

    Purpose: Establish a procedure for beam steering and profile measurement using a 2D ionization chamber array and show equivalence to a water scanning system. Methods: Multiple photon beams (30×30cm{sup 2} field) and electron beams (25×25cm{sup 2} cone) were steered in the radial and transverse directions using Sun Nuclear’s IC PROFILER (ICP). Solid water was added during steering to ensure measurements were beyond the buildup region. With steering complete, servos were zeroed and enabled. Photon profiles were collected in a 30×30cm{sup 2} field at dmax and 2.9 cm depth for flattened and FFF beams respectively. Electron profiles were collected with a 25×25cm{sup 2} cone and effective depth (solid water + 0.9 cm intrinsic buildup) as follows: 0.9 cm (6e), 1.9 cm (9e), 2.9 cm (12e, 16e, 20e). Profiles of the same energy, field size and depth were measured in water with Sun Nuclear’s 3D SCANNER (3DS). Profiles were re-measured using the ICP after the in-water scans. Profiles measured using the ICP and 3DS were compared by (a) examining the differences in Varian’s “Point Difference Symmetry” metric, (b) visual inspection of the overlaid profile shapes and (c) calculation of point-by-point differences. Results: Comparing ICP measurements before and after water scanning showed very good agreement indicating good stability of the linac and measurement system. Comparing ICP Measurements to water phantom measurements using Varian’s symmetry metric showed agreement within 0.5% for all beams. The average magnitude of the agreement was within 0.2%. Comparing ICP Measurements to water phantom measurements using point-by-point difference showed agreement within 0.5% inside of 80% area of the field width. Conclusion: Profile agreement to within 0.5% was observed between ICP and 3DS after steering multiple energies with the ICP. This indicates that the ICP may be used for steering electron beams, and both flattened and FFF photon beams. Song Gao: Sun Nuclear

  4. SU-E-T-460: Impact of the LINAC Repetition Rate On a High-Resolution Liquid Ionization Chamber Array for Patient-Specific QA

    SciTech Connect

    Wang, S; Driewer, J; Zheng, D; Lei, Y; Zhang, Q; Zhu, X; Li, S; Enke, C; Zhou, S; Xu, B

    2015-06-15

    Purpose: The purpose of this study is to investigate the LINAC repetition-rate (dose-rate) dependence of OCTAVIUS 1000SRS liquid ionization chamber (LIC) array for patient specific QA of SRT plans delivered with flattening-filter-free (FFF) beams. Methods: 1) The repetition-rate dependence of 1000SRS was measured in a phantom constructed with 5-cm solid water above and below the array for build-up and backscatter. A 0.3cc calibrated ion chamber was also placed along the central axis 2.3cm below the center chamber of the array for normalizing LINAC output fluctuation. The signals from the center chamber of the array under different repetition rates in the range of 400–2400 MU/min for 6xFFF and 10xFFF beams on a Varian TrueBeamSTx LINAC, normalized by the independent chamber readings, were analyzed for the array response dependence on repetition rates. 2) Twelve Step-and-shoot IMRS QA plans (6xFFF and 10xFFF) were delivered to the array under different repetition rates for analysis and comparison. 3) The absolute doses measured by the center chamber were compared to measurements using an independent ionization chamber with the identical setup, taken as the gold standard. 4) The correction factors based on the actual delivery repetition rate were applied to the measurements, and the results were compared again to the gold standard. Results: 1) The 1000SRS array exhibited repetition-rate dependence for FFF beams, up to 5% for 6xFFF and 10% for 10xFFF; 2) The array showed clinically-acceptable repetition-rate dependence for regular flattened beams; 3) This repetition-rate dependence significantly affected the measurement accuracy, thereby affecting IMRS QA results; 4) By applying an empirical repetition-rate correction, the corrected measurements agreed better with the gold standard ion chamber measurements. Conclusion: OCTAVIUS 1000SRS LIC array exhibited considerable repetition-rate dependence for FFF beams, which will affect the accuracy of the absolute QA

  5. DETECTORS AND EXPERIMENTAL METHODS Study of low momentum track reconstruction for the BESIII main drift chamber

    NASA Astrophysics Data System (ADS)

    Jia, Lu-Kui; Mao, Ze-Pu; Li, Wei-Dong; Cao, Guo-Fu; Cao, Xue-Xiang; Deng, Zi-Yan; He, Kang-Lin; Liu, Chun-Yan; Liu, Huai-Min; Liu, Qiu-Guang; Ma, Qiu-Mei; Ma, Xiang; Qiu, Jin-Fa; Tian, Hao-Lai; Wang, Ji-Ke; Wu, Ling-Hui; Yuan, Ye; Zang, Shi-Lei; Zhang, Chang-Chun; Zhang, Lei; Zhang, Yao; Zhu, Kai; Zou, Jia-Heng

    2010-12-01

    In order to overcome the difficulty brought by the circling charged tracks with transverse momentum less than 120 MeV in the BESIII Main Drift Chamber (MDC), a specialized method called TCurlFinder was developed. This tracking method focuses on the charged track reconstruction under 120 MeV and possesses a special mechanism to reject background noise hits. The performance of the package has been carefully checked and tuned by both Monte Carlo data and real data. The study shows that this tracking method could obviously enhance the reconstruction efficiency in the low transverse momentum region, providing physics analysis with more and reliable data.

  6. Internally heated membrane interfaced to a gas chromatography flame ionization detector.

    PubMed

    Kanu, A Bakarr; Thomas, C L P

    2013-07-15

    Volatile Organic Compounds (VOCs) mixtures in aqueous solutions have been investigated using a simple and efficient all-in-one on-line sampling, isolation, enrichment and pre-concentration internally heated membrane connected to a gas chromatography flame ionization detector (GC-FID). In our previous study with the internally heated membrane, no GC column was used in the instrument. In this new study, we introduce a GC column in the instrument design and this makes it possible for mixtures to be investigated. This new experimental design enabled high resolution separation of analyte mixtures capable of being adsorbed, diffused, and desorbed from the device for detection with an FID. With the new design we present data from investigation of competitive adsorption effects on the membrane. The data showed that analyte adsorption and diffusion onto the membrane can be affected when mixtures of analytes are introduced. The application of multiple linear regressions approach to the data enabled us to correct for the problem of competitive adsorption. Analyte adsorption and diffusion was affected by the diffusion coefficients of the analytes; the higher the diffusion coefficient the better the extraction from the membrane. Increasing the sampling time from 1 to 4 min increases the response by 77%. The sampling time responses were linear up to 4 min sampling time. Above 4 min sampling time, the data deviate from linearity. The effect of adding salt to standards has no effect on analyte absorption onto the membrane. The detection limits for key VOCs studied with an internal standard calibration ranged from 0.2 to 194 ng cm(-3).

  7. Absolute dose measurements by means of a small cylindrical ionization chamber for very high dose per pulse high energy electron beams

    SciTech Connect

    Karaj, E.; Righi, S.; Di Martino, F.

    2007-03-15

    Very high dose per pulse (3-13 cGy/pulse) high energy electron beams are currently produced by special linear accelerators (linac) dedicated to Intra Operative Radiation Therapy (IORT). The electron beams produced by such linacs are collimated by special Perspex applicators of various size and cylindrically shaped. The biggest problems from the dosimetric point of view are caused by the high dose-per-pulse values and the use of inclined applicators. In this work measurements of absolute dose for the inclined applicators were done by using a small cylindrical ionization chamber, type CC01 (Wellhofer), a parallel plane ionization chamber type Markus (PTW 23343) and radiochromic films type EBT. We show a method which allows calculating the quality correction factors for CC01 chamber with an uncertainty of 1% and the absolute dose value for the inclined applicators using CC01 with an uncertainty of 3.1% for electron beams of energy of 6 and 7 MeV produced by the linac dedicated to IORT Novac7.

  8. Simulation of Low Gain Avalanche Detector characteristics based on the concept of negative feedback in irradiated silicon detectors with carrier impact ionization

    NASA Astrophysics Data System (ADS)

    Verbitskaya, E.; Eremin, V.; Zabrodskii, A.; Luukka, P.

    2016-12-01

    In this study the main characteristics of silicon Low Gain Avalanche Detectors (LGAD), the dependencies of the collected charge versus bias voltage and fluence, are calculated to fit experimental data. The calculations are based on two previously developed Ioffe Institute models of radiation degradation in Si detectors: 1) a model of two effective energy levels of radiation-induced defects, and 2) a mechanism of internal negative feedback responsible for the gain degradation in irradiated Si detectors originating from the avalanche multiplication at the detector junction. The combination of these models describes well the properties of irradiated p-i-n detectors in a wide range of fluences. For simulating the LGAD characteristics the models are adapted to its n+-pbi-p-p+ structure, where the built-in boron-doped layer pbi produces high electric field sufficient for carrier impact ionization. It is shown that the developed models give adequate quantitative description of the experimental results for the LGADs up to the fluence of 2×1015 n/cm2 including the detector pulse response; however, additional boron removal from the pbi layer is required to have the best correlation with the experimental data. Similar to the physical model developed for silicon strip detectors operated at high voltage, the results are interpreted in terms of the internal negative feedback mechanism. It is shown that in irradiated LGADs this feedback leads to the transfer of a significant fraction of the potential drop from the built-in layer toward the p+ contact. It initiates two negative effects, which both cause the gain degradation with irradiation: the lowering of the electric field in the n+-pbi region that reduces the multiplication probability, and the increase of the collection time and trapping-related charge losses.

  9. "Smoke": Characterization Of Smoke Particulate For Spacecraft Fire Detection

    NASA Technical Reports Server (NTRS)

    Urban, David L.; Mulholland, George W.; Yang, Jiann; Cleary, Thomas G.; Yuan, Zeng-Guang

    2003-01-01

    The "Smoke" experiment is a flight definition investigation that seeks to increase our understanding of spacecraft fire detection through measurements of particulate size distributions of preignition smokes from typical spacecraft materials. Owing to the catastrophic risk posed by even a very small fire in a spacecraft, the design goal for spacecraft fire detection is to detect the fire as quickly as possible, preferably in the preignition phase before a real flaming fire has developed. Consequently the target smoke for detection is typically not soot (typical of established hydrocarbon fires) but instead, pyrolysis products, and recondensed polymer particles. At the same time, false alarms are extremely costly as the crew and the ground team must respond quickly to every alarm. The U.S. Space Shuttle (STS: Space Transportation System) and the International Space Station (ISS) both use smoke detection as the primary means of fire detection. These two systems were designed in the absence of any data concerning low-gravity smoke particle (and background dust) size distributions. The STS system uses an ionization detector coupled with a sampling pump and the ISS system is a forward light scattering detector operating in the near IR. These two systems have significantly different sensitivities with the ionization detector being most sensitive (on a mass concentration basis) to smaller particulate and the light scattering detector being most sensitive to particulate that is larger than 1 micron. Since any smoke detection system has inherent size sensitivity characteristics, proper design of future smoke detection systems will require an understanding of the background and alarm particle size distributions that can be expected in a space environment.

  10. Physics studies with ICARUS and a hybrid ionization and scintillation fiber detector

    SciTech Connect

    Cline, D.B.

    1992-01-01

    We discuss the physics possibilities for the ICARUS detector currently being tested at CERN. The physics potential goes from a massive proton decay detector to the study of solar neutrinos. In addition, the detection of [nu][sub [mu

  11. Photoneutron production of a Siemens Primus linear accelerator studied by Monte Carlo methods and a paired magnesium and boron coated magnesium ionization chamber system.

    PubMed

    Becker, J; Brunckhorst, E; Schmidt, R

    2007-11-07

    When radiotherapy with photon energies greater than 10 MV is performed neutrons contaminate the photon beam. In this paper the neutron contamination of the 15 MV photon mode of the Siemens Primus accelerator was studied. The Monte Carlo code MCNPX was used for the description of the treatment head and treatment room. The Monte Carlo results were verified by studying the photon depth dose curve and beam profiles in a water phantom. After these verifications the locations of neutron production were studied and the neutron source spectrum and strength were calculated. The neutron response of the paired Mg/Ar and MgB/Ar ionization chamber system was calculated and experimentally verified for two experimental set-ups. The paired chamber system allowed us to measure neutrons inside the field borders and allowed rapid and point wise measurement in contrast to other methods of neutron detection.

  12. Jamming efficiency evaluation of the IR smoke screen against high-orbit IR detector

    NASA Astrophysics Data System (ADS)

    Gao, Gui-qing; Li, Yong-xiang

    2011-08-01

    In order to lower the orientating capability of the DSP satellite, at first the paper analyzes early warning missile satellite detective system, introduces the jamming mechanism of infrared smoke screen, and a model of jamming efficiency evaluation of the IR smoke screen against early warning satellite was built from three sides of absorbency of smoke screen to infrared radiation, dispersion ability and infrared radiation from smoke screen self. At last the correlative conclusion was got based on the brief depiction of Early-warning Satellite.

  13. Instantaneous and continuous measurement of /sup 14/C-labeled substrate oxidation to /sup 14/CO2 by minute tissue specimens: an ionization chamber method

    SciTech Connect

    Davidson, W.D.; Klein, K.L.; Kurokawa, K.; Soll, A.H.

    1981-06-01

    The vibrating reed electrometer and ionization chamber have been adapted for the instantaneous and continuous measurement of /sup 14/C-labeled substrate oxidation to /sup 14/CO2 by minute quantities of isolated tissues. This modified technique, utilizing a ''closed'' circulation incubation system, is 10-50 times as sensitive as the previously described ''open'' circulation techniques. Substrate oxidation curves are described for human erythrocytes and polymorphonuclear leucocytes, canine parietal cells and isolated segments of the rat nephron. This apparatus should prove to be a useful tool for metabolic studies of small quantities of isolated tissue.

  14. (18)F primary standard at ENEA-INMRI by three absolute techniques and calibration of a well-type IG11 ionization chamber.

    PubMed

    Capogni, Marco; Carconi, Pierluigi; De Felice, Pierino; Fazio, Aldo

    2016-03-01

    A new (18)F primary standardization carried out at ENEA-INMRI by three different absolute techniques, i.e. 4πγNaI(Tl)γ high-efficiency counting, TDCR and 4πβ(LS)-γ[NaI(Tl)] coincidence counting method, allowed the calibration of a fixed well-reentrant IG11 ionization chamber (IC), with an uncertainty lower than 1%, and to check the calibration factor of a portable well-type IC NPL-CRC model, previously calibrated. By the new standard the ENEA-INMRI was linked to the BIPM International Reference System (SIR) through the BIPM SIR Transfer Instrument (SIRTI).

  15. Multichannel detectors for profile measurements in clinical proton fields

    SciTech Connect

    Nichiporov, Dmitri; Solberg, Keith; Hsi Wen; Wolanski, Mark; Mascia, Anthony; Farr, Jonathan; Schreuder, Andries

    2007-07-15

    Two beam profile measurement detectors have been developed at Indiana University Cyclotron Facility to address the need for a tool to efficiently verify dose distributions created with active methods of clinical proton beam delivery. The multipad ionization chamber (MPIC) has 128 ionization chambers arranged in one plane and is designed to measure lateral profiles in fields up to 38 cm in diameter. The MPIC pads have a 5 mm pitch for fields up to 20 cm in diameter and a 7 mm pitch for larger fields, providing the accuracy of field size determination about 0.5 mm. The multilayer ionization chamber (MLIC) detector contains 122 small-volume ionization chambers stacked at a 1.82 mm step (water-equivalent) for depth-dose profile measurements. The MLIC detector can measure profiles up to 20 cm in depth, and determine the 80% distal dose fall-off with about 0.1 mm precision. Both detectors can be connected to the same set of electronics modules, which comprise the detectors' data acquisition system. The detectors have been tested in clinical proton fields produced with active methods of beam delivery such as uniform scanning and energy stacking. This article describes detector performance tests and discusses their results. The test results indicate that the MPIC and MLIC detectors can be used for dosimetric characterization of clinical proton fields. The detectors offer significant time savings during measurements in actively delivered beams compared with traditional measurements using a water phantom.

  16. SIGN, a WIMP detector based on high pressure gaseous neon

    NASA Astrophysics Data System (ADS)

    White, J. T.; Gao, J.; Maxin, J.; Miller, J.; Salinas, G.; Wang, H.

    A new WIMP detector concept based on the measurement of Scintillation and Ionization in Gaseous Neon (SIGN) is presented. The detector employs room temperature gaseous neon at a pressure of ≥100 bars as the WIMP target. The ionization is readout using either charge gain or electrofluorescence or both in a modified cylindrical proportional chamber geometry. The primary scintillation is detected by placing a CsI photocathode on the inside wall of the cylindrical chamber. The neon is doped with xenon (≤0.5%) for signal enhancement. Theoretical considerations suggest that the measurement of both scintillation and ionization will provide discrimination between nuclear and electron recoils in this gas mixture.

  17. Quantification of biogenic volatile organic compounds with a flame ionization detector using the effective carbon number concept

    NASA Astrophysics Data System (ADS)

    Faiola, C. L.; Erickson, M. H.; Fricaud, V. L.; Jobson, B. T.; VanReken, T. M.

    2012-08-01

    Biogenic volatile organic compounds (BVOCs) are emitted into the atmosphere by plants and include isoprene, monoterpenes, sesquiterpenes, and their oxygenated derivatives. These BVOCs are among the principal factors influencing the oxidative capacity of the atmosphere in forested regions. BVOC emission rates are often measured by collecting samples onto adsorptive cartridges in the field and then transporting these samples to the laboratory for chromatographic analysis. One of the most commonly used detectors in chromatographic analysis is the flame ionization detector (FID). For quantitative analysis with an FID, relative response factors may be estimated using the effective carbon number (ECN) concept. The purpose of this study was to determine the ECN for a variety of terpenoid compounds to enable improved quantification of BVOC measurements. A dynamic dilution system was developed to make quantitative gas standards of VOCs with mixing ratios from 20-55 ppb. For each experiment using this system, one terpene standard was co-injected with an internal reference, n-octane, and analyzed via an automated cryofocusing system interfaced to a gas chromatograph flame ionization detector and mass spectrometer (GC/MS/FID). The ECNs of 16 compounds (14 BVOCs) were evaluated with this approach, with each test compound analyzed at least three times. The difference between the actual carbon number and measured ECN ranged from -24% to -2%. The difference between theoretical ECN and measured ECN ranged from -22% to 9%. Measured ECN values were within 10% of theoretical ECN values for most terpenoid compounds.

  18. Quantification of biogenic volatile organic compounds with a flame ionization detector using the effective carbon number concept

    NASA Astrophysics Data System (ADS)

    Faiola, C. L.; Erickson, M. H.; Fricaud, V. L.; Jobson, B. T.; VanReken, T. M.

    2012-03-01

    Biogenic volatile organic compounds (BVOCs) are emitted into the atmosphere by plants and include isoprene, monoterpenes, sesquiterpenes, and their oxygenated derivatives. These BVOCs are among the principal factors influencing the oxidative capacity of the atmosphere in forested regions. BVOC emission rates are often measured by collecting samples onto adsorptive cartridges in the field and then transporting these samples to the laboratory for chromatographic analysis. One of the most commonly used detectors in chromatographic analysis is the flame ionization detector (FID). For quantitative analysis with an FID, relative response factors may be estimated using the effective carbon number (ECN) concept. The purpose of this study was to determine the ECN for a variety of terpenoid compounds to enable improved quantification of BVOC measurements. A dynamic dilution system was developed to make quantitative gas standards of VOCs with mixing ratios from 20-55 ppb. For each experiment using this system, one terpene standard was co-injected with an internal reference, n-octane, and analyzed via an automated cryofocusing system interfaced to a gas chromatograph flame ionization detector and mass spectrometer (GC/MS/FID). The ECNs of 16 compounds (14 BVOCs) were evaluated with this approach, with each test compound analyzed at least three times. The difference between the actual carbon number and measured ECN ranged from -24% to -2%. The difference between theoretical ECN and measured ECN ranged from -22% to 9%. Measured ECN values were within 10% of theoretical ECN values for most terpenoid compounds.

  19. The effective point of measurement of ionization chambers and the build-up anomaly in MV x-ray beams.

    PubMed

    McEwen, M R; Kawrakow, I; Ross, C K

    2008-03-01

    A precision experimental investigation of the effective point of measurement (EPOM) of ion chambers in megavoltage beams has been carried out. A one-dimensional scanning phantom system was developed with an overall accuracy in the positioning of a chamber of better than 0.15 mm. Depth-dose data were acquired for a 25 MV beam from an Elekta Precise linac (field sizes of 10 x 10 cm and 25 x 25 cm) for measurement depths in the range 0.6-6 cm. The results confirmed the Monte Carlo calculations of an earlier theoretical investigation by Kawrakow [Med. Phys. 33, 1829-1839 (2006)] that the standard shift for cylindrical chambers, recommended in dosimetry protocols of -0.6r (where r is the internal radius of the cavity), is incorrect. A wide range of ion chambers were investigated and it was found that errors of up to 1.4 mm could occur for certain chamber designs (although typical errors for common chambers were around 0.5 mm). A comparison between measurements and Monte Carlo simulations showed that once the correct EPOM is used, the details of the linac geometry are correct, and the parameters of the electron beam striking the bremsstrahlung target have been adequately determined, the EGSnrc Monte Carlo package is capable of reproducing the experimental data to 0.2 mm or better. The investigation also confirmed that for the highest accuracy depth-dose curves in megavoltage photon beams one should use a well-guarded parallel-plate ion chamber. Three chamber designs were tested here and found to be satisfactory-the Scanditronix-Wellhöfer NACP-02, PTW Roos and Exradin All.

  20. 3D sensitive voxel detector of ionizing radiation based on Timepix device

    NASA Astrophysics Data System (ADS)

    Soukup, P.; Jakubek, J.; Vykydal, Z.

    2011-01-01

    Position sensitive detectors are evolving towards higher segmentation geometries from 0D (single pad) over 1D (strip) to 2D (pixel) detectors. Each step has brought up substantial expansion in the field of applications. The next logical step in this evolution is to design a 3D, i.e. voxel detector. The voxel detector can be constructed from 2D volume element detectors arranged in layers forming a 3D matrix of sensitive elements — voxels. Such detectors can effectively record tracks of energetic particles. By proper analysis of these tracks it is possible to determine the type, direction and energy of the primary particle. One of the prominent applications of such device is in the localization and identification of gamma and neutron sources in the environment. It can be also used for emission and transmission radiography in many fields where standard imagers are currently utilized. The qualitative properties of current imagers such as: spatial resolution, efficiency, directional sensitivity, energy sensitivity and selectivity (background suppression) can be improved. The first prototype of a voxel detector was built using a number of Timepix devices. Timepix is hybrid semiconductor detector consisting of a segmented semiconductor sensor bump-bonded to a readout chip. Each sensor contains 256x256 square pixels of 55 μm size. The voxel detector prototype was successfully tested to prove the concept functionality. The detector has a modular architecture with a daisy chain connection of the individual detector layers. This permits easy rearrangement due to its modularity, while keeping a single readout system for a variable number of detector layers. A limitation of this approach is the relatively large inter-layer distance (4 mm) compared to the pixel thickness (0.3 mm). Therefore the next step in the design is to decrease the space between the 2D detectors.

  1. High performance mini-gas chromatography-flame ionization detector system based on micro gas chromatography column

    NASA Astrophysics Data System (ADS)

    Zhu, Xiaofeng; Sun, Jianhai; Ning, Zhanwu; Zhang, Yanni; Liu, Jinhua

    2016-04-01

    Monitoring Volatile organic compounds (VOCs) was a very important measure for preventing environmental pollution, therefore, a mini gas chromatography (GC) flame ionization detector (FID) system integrated with a mini H2 generator and a micro GC column was developed for environmental VOC monitoring. In addition, the mini H2 generator was able to make the system explode from far away due to the abandoned use of a high pressure H2 source. The experimental result indicates that the fabricated mini GC FID system demonstrated high repeatability and very good linear response, and was able to rapidly monitor complicated environmental VOC samples.

  2. Conception and realization of a parallel-plate free-air ionization chamber for the absolute dosimetry of an ultrasoft X-ray beam

    NASA Astrophysics Data System (ADS)

    Groetz, J.-E.; Ounoughi, N.; Mavon, C.; Belafrites, A.; Fromm, M.

    2014-08-01

    We report the design of a millimeter-sized parallel plate free-air ionization chamber (IC) aimed at determining the absolute air kerma rate of an ultra-soft X-ray beam (E = 1.5 keV). The size of the IC was determined so that the measurement volume satisfies the condition of charged-particle equilibrium. The correction factors necessary to properly measure the absolute kerma using the IC have been established. Particular attention was given to the determination of the effective mean energy for the 1.5 keV photons using the PENELOPE code. Other correction factors were determined by means of computer simulation (COMSOL™and FLUKA). Measurements of air kerma rates under specific operating parameters of the lab-bench X-ray source have been performed at various distances from that source and compared to Monte Carlo calculations. We show that the developed ionization chamber makes it possible to determine accurate photon fluence rates in routine work and will constitute substantial time-savings for future radiobiological experiments based on the use of ultra-soft X-rays.

  3. Conception and realization of a parallel-plate free-air ionization chamber for the absolute dosimetry of an ultrasoft X-ray beam

    SciTech Connect

    Groetz, J.-E. Mavon, C.; Fromm, M.; Ounoughi, N.; Belafrites, A.

    2014-08-15

    We report the design of a millimeter-sized parallel plate free-air ionization chamber (IC) aimed at determining the absolute air kerma rate of an ultra-soft X-ray beam (E = 1.5 keV). The size of the IC was determined so that the measurement volume satisfies the condition of charged-particle equilibrium. The correction factors necessary to properly measure the absolute kerma using the IC have been established. Particular attention was given to the determination of the effective mean energy for the 1.5 keV photons using the PENELOPE code. Other correction factors were determined by means of computer simulation (COMSOL™and FLUKA). Measurements of air kerma rates under specific operating parameters of the lab-bench X-ray source have been performed at various distances from that source and compared to Monte Carlo calculations. We show that the developed ionization chamber makes it possible to determine accurate photon fluence rates in routine work and will constitute substantial time-savings for future radiobiological experiments based on the use of ultra-soft X-rays.

  4. High energy photon reference for radiation protection: technical design of the LINAC beam and ionization chambers; and calculation of monoenergetic conversion coefficients

    NASA Astrophysics Data System (ADS)

    Dusciac, D.; Bordy, J.-M.; Daures, J.; Blideanu, V.

    2016-09-01

    In this work, we present the results of the first part of a research project aimed at offering a complete response to dosimeters providers and nuclear physicists' demands for high-energy (6 - 9 MeV) photon beams for radiation protection purposes. Classical facilities allowing the production of high-energy photonic radiation (proton accelerators, nuclear reactors) are very rare and need large investment for development and use. A novel solution is proposed, consisting in the use of a medical linear accelerator, allowing a significant decrease of all costs.Using Monte Carlo simulations (MCNP5 and PENELOPE codes), a specifically designed electron-photon conversion target allowing for obtaining a high energy photon beam (with an average energy weighted by fluence of about 6 MeV) has been built for radiation protection purposes. Due to the specific design of the target, this "realistic" radiation protection high-energy photon beam presents a uniform distribution of air kerma rate at a distance of 1 m, over a 30 × 30 cm2 surface. Two graphite cavity ionizing chambers for ionometric measurements have been built. For one of these chambers, the charge collection volume has been measured allowing for its use as a primary standard. The second ionizing chamber is used as a transfer standard; as such it has been calibrated in a 60Co beam, and in the high energy photon beam for radiation protection.The measurements with these ionizing chambers allowed for an evaluation of the air kerma rate in the LINAC based high-energy photon beam for radiation protection: the values cover a range between 36 mGy/h and 210 mGy/h, compatible with radiation protection purposes.Finally, using Monte Carlo simulations, conversion coefficients from air kerma to dose equivalent quantities have been calculated in the range between 10 keV and 22.4 MeV, for the spectral distribution of the fluence corresponding to the beam produced by the linear accelerator of the LNE-LNHB.

  5. Physics studies with ICARUS and a hybrid ionization and scintillation fiber detector

    SciTech Connect

    Cline, D.B.

    1992-12-31

    We discuss the physics possibilities for the ICARUS detector currently being tested at CERN. The physics potential goes from a massive proton decay detector to the study of solar neutrinos. In addition, the detection of {nu}{sub {mu}} {yields} {nu}{sub {tau}} and {nu}{sub e} {yields} {nu}{sub {tau}} will be possible with such a detector. One major topic involves the possibility of a complete determination of the MSW solar neutrino parameters with the ICARUS. The possibility of detecting WIMPS with a scintillating fiber liquid Argon (Ar) detector or fiber Xenon (Xe) detector doped with Ar is also described. Some comments on the measurement of the {sup 42}Ar level from an experiment at the Gran Sasso will be made.

  6. Computer program for the sensitivity calculation of a CR-39 detector in a diffusion chamber for radon measurements

    SciTech Connect

    Nikezic, D. Stajic, J. M.; Yu, K. N.

    2014-02-15

    Computer software for calculation of the sensitivity of a CR-39 detector closed in a diffusion chamber to radon is described in this work. The software consists of two programs, both written in the standard Fortran 90 programming language. The physical background and a numerical example are given. Presented software is intended for numerous researches in radon measurement community. Previously published computer programs TRACK-TEST.F90 and TRACK-VISION.F90 [D. Nikezic and K. N. Yu, Comput. Phys. Commun. 174, 160 (2006); D. Nikezic and K. N. Yu, Comput. Phys. Commun. 178, 591 (2008)] are used here as subroutines to calculate the track parameters and to determine whether the track is visible or not, based on the incident angle, impact energy, etching conditions, gray level, and visibility criterion. The results obtained by the software, using five different V functions, were compared with the experimental data found in the literature. Application of two functions in this software reproduced experimental data very well, while other three gave lower sensitivity than experiment.

  7. Track segment finding with CGEM-IT and matching to outer drift chamber tracks in the BESIII detector

    NASA Astrophysics Data System (ADS)

    Sun, Xin-Hua; Wang, Liang-Liang; Wu, Ling-Hui; Ju, Xu-Dong; Xiu, Qing-Lei; Dong, Liao-Yuan; Dong, Ming-Yi; Li, Wei-Dong; Li, Wei-Guo; Liu, Huai-Min; Ou-Yang, Qun; Yuan, Ye; Zhang, Yao

    2016-09-01

    The relative differences in coordinates of Cylindrical Gas Electron Multiplier Detector-based Inner Tracker (CGEM-IT) clusters are studied to search for track segments in CGEM-IT for the BESIII experiment. With the full simulation of single muon track samples, clear patterns are found and parameterized for the correct cluster combinations. The cluster combinations satisfying the patterns are selected as track segment candidates in CGEM-IT with an efficiency higher than 99%. The parameters of the track segments are obtained by a helix fitting. Some χ2 quantities, evaluating the differences in track parameters between the track segments in CGEM-IT and the tracks found in the outer drift chamber, are calculated and used to match them. Proper χ2 requirements are determined as a function of transverse momentum and the matching efficiency is found to be reasonable. Supported by National Key Basic Research Program of China (2015CB856706), National Natural Science Foundation of China (11575222, 11205184, 11205182, 11121092, 11475185) and Joint Funds of National Natural Science Foundation of China (U1232201)

  8. Computer program for the sensitivity calculation of a CR-39 detector in a diffusion chamber for radon measurements

    NASA Astrophysics Data System (ADS)

    Nikezic, D.; Yu, K. N.; Stajic, J. M.

    2014-02-01

    Computer software for calculation of the sensitivity of a CR-39 detector closed in a diffusion chamber to radon is described in this work. The software consists of two programs, both written in the standard Fortran 90 programming language. The physical background and a numerical example are given. Presented software is intended for numerous researches in radon measurement community. Previously published computer programs TRACK_TEST.F90 and TRACK_VISION.F90 [D. Nikezic and K. N. Yu, Comput. Phys. Commun. 174, 160 (2006); D. Nikezic and K. N. Yu, Comput. Phys. Commun. 178, 591 (2008)] are used here as subroutines to calculate the track parameters and to determine whether the track is visible or not, based on the incident angle, impact energy, etching conditions, gray level, and visibility criterion. The results obtained by the software, using five different V functions, were compared with the experimental data found in the literature. Application of two functions in this software reproduced experimental data very well, while other three gave lower sensitivity than experiment.

  9. Comparison of ionisation chamber and semiconductor detector devices for measurement of the dose-width product for panoramic dental units.

    PubMed

    Mitchell, S A; Martin, C J

    2013-06-01

    Doses for panoramic dental radiography are assessed in terms of the dose-width product (DWP) or dose-area product, which gives a measure of the radiation through a whole exposure. The DWP can be measured using a pencil ionisation chamber (IC) similar to that used for computed tomography dose assessment. However, ICs are sensitive to radiation incident from all directions and so backscatter from the image receptor may increase the recorded dose. This study compares measurements performed using four options: a pencil IC mounted straight on the image receptor, the IC mounted with a steel plate to the rear to standardise scatter conditions, the IC mounted with a steel plate and lead collimators in front to minimise the effect of extra-focal radiation, and a Quart Dido employing a one square centimetre semiconductor detector (SD) designed for panoramic measurements. The results indicate that modification of the current method by incorporating a steel plate reduced the measurement dose by 7% on average, but the reduction was greater for units with semiconductor imaging plates. The measurements with the SD agree more closely with the IC with the steel plate to the rear. An IC with a backing plate to standardise scatter or a suitable SD is recommended for measurement on panoramic dental units.

  10. Correction factors for the NMi free-air ionization chamber for medium-energy x-rays calculated with the Monte Carlo method.

    PubMed

    Grimbergen, T W; van Dijk, E; de Vries, W

    1998-11-01

    A new method is described for the determination of x-ray quality dependent correction factors for free-air ionization chambers. The method is based on weighting correction factors for mono-energetic photons, which are calculated using the Monte Carlo method, with measured air kerma spectra. With this method, correction factors for electron loss, scatter inside the chamber and transmission through the diaphragm and front wall have been calculated for the NMi free-air chamber for medium-energy x-rays for a wide range of x-ray qualities in use at NMi. The newly obtained correction factors were compared with the values in use at present, which are based on interpolation of experimental data for a specific set of x-ray qualities. For x-ray qualities which are similar to this specific set, the agreement between the correction factors determined with the new method and those based on the experimental data is better than 0.1%, except for heavily filtered x-rays generated at 250 kV. For x-ray qualities dissimilar to the specific set, differences up to 0.4% exist, which can be explained by uncertainties in the interpolation procedure of the experimental data. Since the new method does not depend on experimental data for a specific set of x-ray qualities, the new method allows for a more flexible use of the free-air chamber as a primary standard for air kerma for any x-ray quality in the medium-energy x-ray range.

  11. SU-E-P-35: Real-Time Patient Transit Dose Verification of Volumetric Modulated Arc Radiotherapy by a 2D Ionization Chamber Array

    SciTech Connect

    Liu, X

    2015-06-15

    Purpose: To explore the real-time dose verification method in volumetric modulated arc radiotherapy (VMAT) with a 2D array ion chamber array. Methods: The 2D ion chamber array was fixed on the panel of electronic portal imaging device (EPID). Source-detector distance (SDD)was 140cm. 8mm RW3 solid water was added to the detector panel to achieve maximum readings.The patient plans for esophageal, prostate and liver cancers were selected to deliver on the cylindrical Cheese phantom 5 times in order to validate the reproducibility of doses. Real-time patient transit dose measurements were performed at each fraction. Dose distributions wereevaluated using gamma index criteria of 3mm DTA and 3% dose difference referred to the firsttime Result. Results: The gamma index pass rate in the Cheese phantom were about 98%; The gamma index pass rate for esophageal, liver and prostate cancer patient were about 92%,94%, and 92%, respectively; Gamma pass rate for all single fraction were more than 90%. Conclusion: The 2D array is capable of monitoring the real time transit doses during VMAT delivery. It is helpful to improve the treatment accuracy.

  12. Gated strip proportional detector

    DOEpatents

    Morris, Christopher L.; Idzorek, George C.; Atencio, Leroy G.

    1987-01-01

    A gated strip proportional detector includes a gas tight chamber which encloses a solid ground plane, a wire anode plane, a wire gating plane, and a multiconductor cathode plane. The anode plane amplifies the amount of charge deposited in the chamber by a factor of up to 10.sup.6. The gating plane allows only charge within a narrow strip to reach the cathode. The cathode plane collects the charge allowed to pass through the gating plane on a set of conductors perpendicular to the open-gated region. By scanning the open-gated region across the chamber and reading out the charge collected on the cathode conductors after a suitable integration time for each location of the gate, a two-dimensional image of the intensity of the ionizing radiation incident on the detector can be made.

  13. Gated strip proportional detector

    DOEpatents

    Morris, C.L.; Idzorek, G.C.; Atencio, L.G.

    1985-02-19

    A gated strip proportional detector includes a gas tight chamber which encloses a solid ground plane, a wire anode plane, a wire gating plane, and a multiconductor cathode plane. The anode plane amplifies the amount of charge deposited in the chamber by a factor of up to 10/sup 6/. The gating plane allows only charge within a narrow strip to reach the cathode. The cathode plane collects the charge allowed to pass through the gating plane on a set of conductors perpendicular to the open-gated region. By scanning the open-gated region across the chamber and reading out the charge collected on the cathode conductors after a suitable integration time for each location of the gate, a two-dimensional image of the intensity of the ionizing radiation incident on the detector can be made.

  14. Output current variation and polarity effect by electric field and ion-pair non-uniformity inside thimble-type ionization chamber

    NASA Astrophysics Data System (ADS)

    Kim, Jaecheon; Kim, Yong Kyun; Kim, Soon Young; Kim, Jong Kyung

    2007-09-01

    A new analytic approach considering both electric field and ion-pair non-uniformity has been proposed to accurately analyze the design characteristics of an ionization chamber and to interpret measurements. It is commonly assumed that ion-pairs are generated uniformly in the air volume, but such an assumption ignores various source and geometry conditions. The new approach was applied to angular dependence analysis and to polarity effect assessment in an ionization chamber. For the angular dependence analysis, whole, uniform, and non-uniform output currents were calculated as a function of the irradiation angle for an 241Am gamma-ray source. The non-uniform output current proposed in this paper was found to be closer to the measured one. This is because the non-uniform output current takes into account the ion-pair distribution in the air volume as well as the active volume determined by the electric field. For the polarity effect assessment, the amount of field distortion due to potential difference and actual current difference was calculated. Previous methods cannot appropriately estimate the variation of polarity effect because they ignore the influence of the ion-pair distribution. The polarity effect assessment using the non-uniform output current can be more useful for obtaining the practical current difference, because this assessment considers both the variation of active volume and the ion-pair non-uniformity according to source conditions such as the irradiation angle and the distance. It is important to precisely calculate not only the active volume, but also the variation in the ion-pair distribution.

  15. Reference dosimetry at the Australian Synchrotron's imaging and medical beamline using free-air ionization chamber measurements and theoretical predictions of air kerma rate and half value layer

    SciTech Connect

    Crosbie, Jeffrey C.; Rogers, Peter A. W.; Stevenson, Andrew W.; Hall, Christopher J.; Lye, Jessica E.; Nordstroem, Terese; Midgley, Stewart M.; Lewis, Robert A.

    2013-06-15

    Purpose: Novel, preclinical radiotherapy modalities are being developed at synchrotrons around the world, most notably stereotactic synchrotron radiation therapy and microbeam radiotherapy at the European Synchrotron Radiation Facility in Grenoble, France. The imaging and medical beamline (IMBL) at the Australian Synchrotron has recently become available for preclinical radiotherapy and imaging research with clinical trials, a distinct possibility in the coming years. The aim of this present study was to accurately characterize the synchrotron-generated x-ray beam for the purposes of air kerma-based absolute dosimetry. Methods: The authors used a theoretical model of the energy spectrum from the wiggler source and validated this model by comparing the transmission through copper absorbers (0.1-3.0 mm) against real measurements conducted at the beamline. The authors used a low energy free air ionization chamber (LEFAC) from the Australian Radiation Protection and Nuclear Safety Agency and a commercially available free air chamber (ADC-105) for the measurements. The dimensions of these two chambers are different from one another requiring careful consideration of correction factors. Results: Measured and calculated half value layer (HVL) and air kerma rates differed by less than 3% for the LEFAC when the ion chamber readings were corrected for electron energy loss and ion recombination. The agreement between measured and predicted air kerma rates was less satisfactory for the ADC-105 chamber, however. The LEFAC and ADC measurements produced a first half value layer of 0.405 {+-} 0.015 and 0.412 {+-} 0.016 mm Cu, respectively, compared to the theoretical prediction of 0.427 {+-} 0.012 mm Cu. The theoretical model based upon a spectrum calculator derived a mean beam energy of 61.4 keV with a first half value layer of approximately 30 mm in water. Conclusions: The authors showed in this study their ability to verify the predicted air kerma rate and x-ray attenuation

  16. Results from the Medipix-2 and Timepix Ionizing Radiation Imaging Detectors on BEXUS stratospheric balloon student campaigns

    NASA Astrophysics Data System (ADS)

    Urbar, Jaroslav; Scheirich, Jan; Jakubek, Jan

    Results of the first two experiments using the semiconductor pixel detectors of the Medipix fam-ily for energetic particle imaging in the stratospheric environment are presented. The original detecting device was based on the hybrid pixel detectors of Medipix-2 and Timepix developed at CERN with USB interface developed at Institute of Experimental and Applied Physics of Czech Technical University in Prague. The detectors were used in tracking mode allowing them to operate as an active nuclear emulsion". The actual flight time of BEXUS7 with Medipix-2 on 8th October 2008 was over 4 hours, with 2 hours at stable floating altitude of 26km. BEXUS9 measurements of 3.5 hour duration by Timepix, Medipix-2 and ST-6 Geiger telescope instruments took place in arctic atmosphere till ceiling altitude of 24km on 11th October 2009. Stratospheric balloon platform is the optimal realization for all in-situ measurements of atmo-spheric electricity. Not only because of the high altitudes reached, but also due to its slow ascent velocity for statistically relevant sampling of the ambient environment for improving cosmic ray induced ionisation rate model inputs. The flight opportunity for BEXUS student projects was provided by Education department of the European Space Agency (ESA) and Eu-rolaunch -Collaboration of Swedish National Space Board (SNSB) and German Space Agency (DLR). The scientific goal was to check energetic particle type altitudinal dependencies, si-multaneously testing proper detector calibration by detecting fluxes of ionizing radiation while evaluating instrumentation endurance and performance.

  17. Development of a wide-range tritium-concentration detector

    SciTech Connect

    Jun, F.; Zhe, L.; Shicheng, L.; Jiangfeng, S.; Deli, L.

    2015-03-15

    According to the requirements of the tritium related systems of the TBM (Test Blanket Module) for monitoring the on-line tritium concentration, a wide-range tritium-concentration detector has been developed to measure the tritium concentration in the range of 10{sup 4} Bq/ml - 5*10{sup 8} Bq/ml. This detector is combined with a low-memory helium ionization chamber. The weak current signal collected in the ionization chamber is converted to the voltage signal by an I-V converter. The minimum weak current which the detector could be measured is 10{sup -14} A. The performance of the background current and the current response linearity of the prototype have been tested. The test result indicates that the linear response of the current signal of the prototype without connecting the ionization chamber is good. The linear correlation coefficient is R{sup 2} = 0.998.

  18. Rapid and simultaneous analysis of ten aromatic amines in mainstream cigarette smoke by liquid chromatography/electrospray ionization tandem mass spectrometry under ISO and "Health Canada intensive" machine smoking regimens.

    PubMed

    Xie, Fuwei; Yu, Jingjing; Wang, Sheng; Zhao, Ge; Xia, Qiaoling; Zhang, Xiaobing; Zhang, Shusheng

    2013-10-15

    Ten primary aromatic amines (AAs) in mainstream cigarette smoke under both ISO and "Health Canada intensive" machine smoking regimens were determined in this work, which were suspected to be carcinogenic compounds. The measured AAs included aniline, ortho-toluidine, meta-toluidine, para-toluidine, 1-naphthylamine, 2-naphthylamine, 3-aminobiphenyl, 4-aminobiphenyl, meta-phenylenediamine and meta-anisidine. For rapidly and sensitively analyzing these AAs, a liquid chromatography-electrospray ionization tandem mass spectrometric (LC-MS/MS) method coupled with solid phase extraction (SPE) was developed. The particulate phase of mainstream cigarette smoke was collected on a Cambridge filter pads, while the gas phase was trapped by 25 mL 5% HCl solution. Then, the pad was extracted in an ultrasonic bath with the impinger HCl solution. After being neutralized with NaOH, the extract was purified with a HLB solid phase extraction column, and then was analyzed with LC-MS/MS using isotope-labeled internal standard. The overall sample pretreatment and analysis time was less than 1.5h. The limits of detection for all targets ranged from 0.05 ng cig(-1) to 0.96 ng cig(-1) with the recoveries in the range of 75.0-131.8%. And the intra-day and inter-day precisions were less than 10% and 16%, respectively. Under HCI machine smoking regimen, the AAs yields in mainstream cigarette smoke were much higher and the average increases were greater than 100% compared with those under ISO smoking condition.

  19. Measurements of miniature ionization chamber currents in the JSI TRIGA Mark II reactor demonstrate the importance of the delayed contribution to the photon field in nuclear reactors

    NASA Astrophysics Data System (ADS)

    Radulović, Vladimir; Fourmentel, Damien; Barbot, Loïc; Villard, Jean-François; Kaiba, Tanja; Gašper, Žerovnik; Snoj, Luka

    2015-12-01

    The characterization of experimental locations of a research nuclear reactor implies the determination of neutron and photon flux levels within, with the best achievable accuracy. In nuclear reactors, photon fluxes are commonly calculated by Monte Carlo simulations but rarely measured on-line. In this context, experiments were conducted with a miniature gas ionization chamber (MIC) based on miniature fission chamber mechanical parts, recently developed by the CEA (French Atomic Energy and Alternative Energies Commission) irradiated in the core of the Jožef Stefan Institute TRIGA Mark II reactor in Ljubljana, Slovenia. The aim of the study was to compare the measured MIC currents with calculated currents based on simulations with the MCNP6 code. A discrepancy of around 50% was observed between the measured and the calculated currents; in the latter taking into consideration only the prompt photon field. Further experimental measurements of MIC currents following reactor SCRAMs (reactor shutdown with rapid insertions of control rods) provide evidence that over 30% of the total measured signal is due to the delayed photon field, originating from fission and activation products, which are untreated in the calculations. In the comparison between the measured and calculated values, these findings imply an overall discrepancy of less than 20% of the total signal which is still unexplained.

  20. Assaying multiple 125I seeds with the well-ionization chamber SourceCheck4π 33005 and a new insert

    PubMed Central

    Ballester, Facundo; Perez-Calatayud, Jose; Vijande, Javier

    2015-01-01

    Purpose To provide a practical solution that can be adopted in clinical routine to fulfill the AAPM-ESTRO recommendations regarding quality assurance of seeds used in prostate permanent brachytherapy. The aim is to design a new insert for the well-ionization chamber SourceCheck4π 33005 (PTW, Germany) that allows evaluating the mean air-kerma strength of up to ten 125I seeds with one single measurement instead of measuring each seed individually. Material and methods The material required is: a) the SourceCheck4π 33005 well-ionization chamber provided with a PTW insert to measure the air-kerma strength S K of one single seed at a time; b) a newly designed insert that accommodates ten seeds in one column, which allows measuring the mean S K of the ten seeds in one single measurement; and c) a container with ten seeds from the same batch and class of the seeds used for the patient implant, and a set of nine non-radioactive seeds. The new insert is characterized by determining its calibration coefficient, used to convert the reading of the well-chamber when ten seeds are measured to their mean S K. The proposed method is validated by comparing the mean S K of the ten seeds obtained from the new insert with the individual measurement of S K of each seed, evaluated with the PTW insert. Results The ratio between the calibration coefficient of the new insert and the calibration coefficient of the PTW insert for the SourceCheck4π 33005 is 1.135 ± 0.007 (k = 1). The mean S K of a set of ten seeds evaluated with this new system is in agreement with the mean value obtained from measuring independently the S K of each seed. Conclusions The new insert and procedure allow evaluating the mean S K of ten seeds prior to the implant in a single measurement. The method is faster and more efficient from radiation protection point of view than measuring the individual S K of each seed. PMID:26816507

  1. Computed tomography dosimetry with high-resolution detectors commonly used in radiotherapy - an energy dependence study.

    PubMed

    Liebmann, Mario; Poppe, Bjoern; von Boetticher, Heiner

    2015-09-08

    New methods of dosimetry in computed tomography (CT) X-ray fields require the use of high-resolution detectors instead of pencil-type ionization chambers typically used for CT dose index (CTDI) measurements. This paper presents a study on the suitability of a wide range of ionization chambers, diodes, and a two-dimensional detector array, used primarily in radiation therapy, for CT and cone-beam CT dosimetry. Specifically, the energy dependence of these detectors from 50 kVp up to 125 kVp is reported. All measurements were performed in reference to a calibrated diode for use in this energy region. The radiation quality correction factors provided by the manufacturer were used, depending on the measured half-value layer (HVL) for the particular X-ray beam. Our study demonstrated the general usability of thimble ionization chambers. These thimble ionization chambers showed a maximum variation in energy response of 5%. Ionization chambers with even smaller sensitive volume, and which exhibit similar variation in energy dependence, can be used if higher spatial resolution is required. Furthermore, the investigated detectors are better suited for dosimetry at CT and CBCT units than conventional large volume or flat detectors, due to their rotational symmetry. Nevertheless, a flat detector can be used for certain measurement tasks, such as the acquisition of percent depth-dose curves or beam profiles for nonrotating beams, which are important for beam characterization.

  2. Gating characteristics of an Elekta radiotherapy treatment unit measured with three types of detector.

    PubMed

    Evans, Philip M; Symonds-Tayler, J Richard N; Colgan, Ruth; Hugo, Geoffrey D; Letts, Nicky; Sandin, Carlos

    2010-04-21

    The characteristics of an Elekta Precise treatment machine with a gating interface were investigated. Three detectors were used: a Farmer ionization chamber, a MatriXX ionization chamber array and an in-house, single pulse-measurement ionization chamber (IVC). Measurements were made of dosimetric accuracy, flatness and symmetry characteristics and duty cycle for a range of beam-on times and gating periods. Results were compared with a standard ungated delivery as a reference. For all beam-on times, down to 0.5 s, dosimetric differences were below +/-1% and flatness and symmetry parameter variations were below +/-1.5%. For the shorter beam-on times the in-house detector deviated from the other two detectors, suggesting that this device should be used in conjunction with other detectors for absolute dosimetry purposes. However, it was found to be useful for studying gated beam characteristics pulse by pulse.

  3. NOTE: Gating characteristics of an Elekta radiotherapy treatment unit measured with three types of detector

    NASA Astrophysics Data System (ADS)

    Evans, Philip M.; Symonds-Tayler, J. Richard N.; Colgan, Ruth; Hugo, Geoffrey D.; Letts, Nicky; Sandin, Carlos

    2010-04-01

    The characteristics of an Elekta Precise treatment machine with a gating interface were investigated. Three detectors were used: a Farmer ionization chamber, a MatriXX ionization chamber array and an in-house, single pulse-measurement ionization chamber (IVC). Measurements were made of dosimetric accuracy, flatness and symmetry characteristics and duty cycle for a range of beam-on times and gating periods. Results were compared with a standard ungated delivery as a reference. For all beam-on times, down to 0.5 s, dosimetric differences were below ±1% and flatness and symmetry parameter variations were below ±1.5%. For the shorter beam-on times the in-house detector deviated from the other two detectors, suggesting that this device should be used in conjunction with other detectors for absolute dosimetry purposes. However, it was found to be useful for studying gated beam characteristics pulse by pulse.

  4. Penning Ionization Electron Spectroscopy in Glow Discharge: A New Dimension for Gas Chromatography Detectors

    NASA Technical Reports Server (NTRS)

    Sheverev, V. A.; Khromov, N. A.; Kojiro, D. R.; Fonda, Mark (Technical Monitor)

    2002-01-01

    Admixtures to helium of 100 ppm and 5 ppm of nitrogen, and 100 ppm and 10 ppm of carbon monoxide were identified and measured in the helium discharge afterglow using an electrical probe placed into the plasma. For nitrogen and carbon monoxide gases, the measured electron energy spectra display distinct characteristic peaks (fingerprints). Location of the peaks on the energy scale is determined by the ionization energies of the analyte molecules. Nitrogen and carbon monoxide fingerprints were also observed in a binary mixture of these gases in helium, and the relative concentration analytes has been predicted. The technically simple and durable method is considered a good candidate for a number of analytical applications, and in particular, in GC and for analytical flight instrumentation.

  5. New structural limits on magma chamber locations at the Valley of Ten Thousand Smokes, Katmai National Park, Alaska

    NASA Astrophysics Data System (ADS)

    Wallmann, Peter C.; Pollard, David D.; Hildreth, Wes; Eichelberger, John C.

    1990-12-01

    New structural data from the Novarupta basin, Katmai National Park, Alaska, site of the largest volcanic eruption of this century (1912), provide limits for the location of magma chambers associated with this eruption. To investigate the subsurface structure of the 1912 vents, and to support an interdisciplinary study of this young volcanic system, a project of geologic mapping of surficial and bedrock structures in the vent region of the 1912 eruption has been undertaken.Landslide scarps, arcuate grabens, a monoclinal fold, and truncated ridges circumscribe the Novarupta basin, marking the inferred outer rim of the vent. A set of radial fissures crosses the southern margin of the basin, striking ˜140°, subparallel to the dominant bedrock joint set. These fissures and joints, along with the local plate-motion vector and the inferred regional stress orientation, are consistent with a feeder dike propagating from a reservoir beneath Trident volcano to the eruptive vent.

  6. New structural limits on magma chamber locations at the Valley of Ten Thousand Smokes, Katmai National Park, Alaska

    SciTech Connect

    Wallmann, P.C.; Pollard, D.D. ); Hildreth, W. ); Eichelberger, J.C. )

    1990-12-01

    New structural data from the Novarupta basin, Katmai National Park, Alaska, site of the largest volcanic eruption of this century (1912), provide limits for the location of magma chambers associated with this eruption. To investigate the subsurface structure of the 1912 vents, and to support an interdisciplinary study of this young volcanic system, a project of geologic mapping of surficial and bedrock structures in the vent region of the 1912 eruption has been undertaken. Landslide scarps, arcuate grabens, a monoclinal fold, and truncated ridges circumscribe the Novarupta basin, marking the inferred outer rim of the vent. A set of radial fissures crosses the southern margin of the basin, striking {approximately}140{degree}, subparallel to the dominant bedrock joint set. These fissures and joints, along with the local plate-motion vector and the inferred regional stress orientation, are consistent with a feeder dike propagating from a reservoir beneath Trident volcano to the eruptive vent.

  7. Examination system utilizing ionizing radiation and a flexible, miniature radiation detector probe

    DOEpatents

    Majewski, Stanislaw; Kross, Brian J.; Zorn, Carl J.; Majewski, Lukasz A.

    1996-01-01

    An optimized examination system and method based on the Reverse Geometry X-Ray.RTM. (RGX.RTM.) radiography technique are presented. The examination system comprises a radiation source, at least one flexible, miniature radiation detector probe positioned in appropriate proximity to the object to be examined and to the radiation source with the object located between the source and the probe, a photodetector device attachable to an end of the miniature radiation probe, and a control unit integrated with a display device connected to the photodetector device. The miniature radiation detector probe comprises a scintillation element, a flexible light guide having a first end optically coupled to the scintillation element and having a second end attachable to the photodetector device, and an opaque, environmentally-resistant sheath surrounding the flexible light guide. The probe may be portable and insertable, or may be fixed in place within the object to be examined. An enclosed, flexible, liquid light guide is also presented, which comprises a thin-walled flexible tube, a liquid, preferably mineral oil, contained within the tube, a scintillation element located at a first end of the tube, closures located at both ends of the tube, and an opaque, environmentally-resistant sheath surrounding the flexible tube. The examination system and method have applications in non-destructive material testing for voids, cracks, and corrosion, and may be used in areas containing hazardous materials. In addition, the system and method have applications for medical and dental imaging.

  8. Examination system utilizing ionizing radiation and a flexible, miniature radiation detector probe

    DOEpatents

    Majewski, S.; Kross, B.J.; Zorn, C.J.; Majewski, L.A.

    1996-10-22

    An optimized examination system and method based on the Reverse Geometry X-Ray{trademark} (RGX{trademark}) radiography technique are presented. The examination system comprises a radiation source, at least one flexible, miniature radiation detector probe positioned in appropriate proximity to the object to be examined and to the radiation source with the object located between the source and the probe, a photodetector device attachable to an end of the miniature radiation probe, and a control unit integrated with a display device connected to the photodetector device. The miniature radiation detector probe comprises a scintillation element, a flexible light guide having a first end optically coupled to the scintillation element and having a second end attachable to the photodetector device, and an opaque, environmentally-resistant sheath surrounding the flexible light guide. The probe may be portable and insertable, or may be fixed in place within the object to be examined. An enclosed, flexible, liquid light guide is also presented, which comprises a thin-walled flexible tube, a liquid, preferably mineral oil, contained within the tube, a scintillation element located at a first end of the tube, closures located at both ends of the tube, and an opaque, environmentally-resistant sheath surrounding the flexible tube. The examination system and method have applications in non-destructive material testing for voids, cracks, and corrosion, and may be used in areas containing hazardous materials. In addition, the system and method have applications for medical and dental imaging. 5 figs.

  9. Quantification of Biogenic Volatile Organic Compounds with a Flame Ionization Detector Using the Effective Carbon Number Concept

    NASA Astrophysics Data System (ADS)

    Faiola, C. L.; Erickson, M. H.; Fricaud, V. L.; Wallace, H. W.; Jobson, B. T.; VanReken, T. M.

    2011-12-01

    Biogenic volatile organic compounds (BVOCs) are emitted into the atmosphere by plants and include structurally complex organic molecules such as monoterpenes, sesquiterpenes, and their oxygenated derivatives. These BVOCs are among the principle factors influencing the oxidative capacity of the atmosphere in forested regions. BVOC emission rates are often measured by collecting samples onto adsorptive cartridges in the field and then transporting these samples to the laboratory for chromatographic analysis. One of the most commonly used detectors in gas chromatography is the flame ionization detector (FID) due to its broad linear range, high sensitivity, and predictable response to many compounds. The FID response to saturated aliphatic molecules is proportional to carbon number. However, deviations occur as the complexity of the molecular structure increases. To account for these deviations, Sternberg et al. (1962) developed the effective carbon number (ECN) concept, which describes the number of carbons the FID "effectively" responds to. The ECN of a complex molecule can be estimated from the number and type of functional groups present, allowing an estimate of relative response factors for quantification. This approach is particularly useful for applications where samples contain a mixture of organic compounds and standards are not realistically accessible- a common predicament for environmental measurements. ECNs for a limited number of compounds with varying functional groups have been quantified in previous studies. However, there remain large gaps in the variety of compounds for which published data are available. This results in higher than necessary uncertainties when quantifying compounds that are structurally dissimilar to those that have been reported in the literature. The purpose of this study was to determine the ECN for a variety of terpenoid compounds to enable improved quantification of BVOC measurements. A dynamic dilution system was developed to

  10. Detectors

    DOEpatents

    Orr, Christopher Henry; Luff, Craig Janson; Dockray, Thomas; Macarthur, Duncan Whittemore; Bounds, John Alan; Allander, Krag

    2002-01-01

    The apparatus and method provide techniques through which both alpha and beta emission determinations can be made simultaneously using a simple detector structure. The technique uses a beta detector covered in an electrically conducting material, the electrically conducting material discharging ions generated by alpha emissions, and as a consequence providing a measure of those alpha emissions. The technique also offers improved mountings for alpha detectors and other forms of detectors against vibration and the consequential effects vibration has on measurement accuracy.

  11. DETECTORS AND EXPERIMENTAL METHODS: Performance testing of a long-strip two-end readout multi-gap resistive plate chamber

    NASA Astrophysics Data System (ADS)

    Wang, Xiao-Bin; Wang, Yi; Luo, Ming; Li, Yuan-Jing; Cheng, Jian-Ping

    2009-02-01

    Multi-gap Resistive Plate Chamber (MRPC) is a new generation of gas detector with good timing and spacial resolution, whose technique is widely applied in some recent high energy (nuclear) physics experiments. In this letter, we report a long-strip two-end readout MRPC and its test beam performance. The measurements show that the long-strip performs a transmission line characteristic and the impedance is independent of the length of strip. The MRPC module we developed is presented to gain a timing resolution of ~80 ps and a spacial resolution of ~6.4 mm. The possible application of the MRPC is also discussed.

  12. CONTINUOUS ROTATION SCATTERING CHAMBER

    DOEpatents

    Verba, J.W.; Hawrylak, R.A.

    1963-08-01

    An evacuated scattering chamber for use in observing nuclear reaction products produced therein over a wide range of scattering angles from an incoming horizontal beam that bombards a target in the chamber is described. A helically moving member that couples the chamber to a detector permits a rapid and broad change of observation angles without breaching the vacuum in the chamber. Also, small inlet and outlet openings are provided whose size remains substantially constant. (auth)

  13. Xenon bubble chambers for direct dark matter detection

    NASA Astrophysics Data System (ADS)

    Levy, C.; Fallon, S.; Genovesi, J.; Khaitan, D.; Klimov, K.; Mock, J.; Szydagis, M.

    2016-03-01

    The search for dark matter is one of today's most exciting fields. As bigger detectors are being built to increase their sensitivity, background reduction is an ever more challenging issue. To this end, a new type of dark matter detector is proposed, a xenon bubble chamber, which would combine the strengths of liquid xenon TPCs, namely event by event energy resolution, with those of a bubble chamber, namely insensitivity to electronic recoils. In addition, it would be the first time ever that a dark matter detector is active on all three detection channels, ionization and scintillation characteristic of xenon detectors, and heat through bubble formation in superheated fluids. Preliminary simulations show that, depending on threshold, a discrimination of 99.99% to 99.9999+% can be achieved, which is on par or better than many current experiments. A prototype is being built at the University at Albany, SUNY. The prototype is currently undergoing seals, thermal, and compression testing.

  14. The eye (and brain) as ionizing particle detector? First results from the ALTEA - space experiment

    NASA Astrophysics Data System (ADS)

    Narici, Livio

    The first part of ALTEA-Space experiments have been performed on the ISS (USLab) between August 2006 and July 2007. The ALTEA hardware features 6 particle telescopes each with 6 striped 8 x 16 cm2 silicon planes arranged alternately in the x and y direction. These detectors are hold by helmet shaped holder. ALTEA features also a light tight visual stimulation unit, a 32 channel EEG cap and electronics, a 3-buttons pushbutton. Two different experiment modalities were run: DOSI and CNSM. The former is the study of the radiation environment of the USLab, and results from these measurements are mostly covered by other papers in this conference; the latter is the study of the electrophysiological activity in coincidence with particle traveling through the eye/brain of the astronaut, with specific reference to the observed light flashes. In this paper we will present first results from these measurements and discuss, within this panorama, the amount of the measured radiation in the brain/eye. Seven CNSM sessions have been performed (on three astronauts), with a total of 20 light flashes perceived. Comparisons with previous measurements in Low Earth Orbit and during the flights to the Moon will be also shown

  15. An Efficient, FPGA-Based, Cluster Detection Algorithm Implementation for a Strip Detector Readout System in a Time Projection Chamber Polarimeter

    NASA Technical Reports Server (NTRS)

    Gregory, Kyle J.; Hill, Joanne E. (Editor); Black, J. Kevin; Baumgartner, Wayne H.; Jahoda, Keith

    2016-01-01

    A fundamental challenge in a spaceborne application of a gas-based Time Projection Chamber (TPC) for observation of X-ray polarization is handling the large amount of data collected. The TPC polarimeter described uses the APV-25 Application Specific Integrated Circuit (ASIC) to readout a strip detector. Two dimensional photoelectron track images are created with a time projection technique and used to determine the polarization of the incident X-rays. The detector produces a 128x30 pixel image per photon interaction with each pixel registering 12 bits of collected charge. This creates challenging requirements for data storage and downlink bandwidth with only a modest incidence of photons and can have a significant impact on the overall mission cost. An approach is described for locating and isolating the photoelectron track within the detector image, yielding a much smaller data product, typically between 8x8 pixels and 20x20 pixels. This approach is implemented using a Microsemi RT-ProASIC3-3000 Field-Programmable Gate Array (FPGA), clocked at 20 MHz and utilizing 10.7k logic gates (14% of FPGA), 20 Block RAMs (17% of FPGA), and no external RAM. Results will be presented, demonstrating successful photoelectron track cluster detection with minimal impact to detector dead-time.

  16. Wide range radioactive gas concentration detector

    DOEpatents

    Anderson, David F.

    1984-01-01

    A wide range radioactive gas concentration detector and monitor which is capable of measuring radioactive gas concentrations over a range of eight orders of magnitude. The device of the present invention is designed to have an ionization chamber which is sufficiently small to give a fast response time for measuring radioactive gases but sufficiently large to provide accurate readings at low concentration levels. Closely spaced parallel plate grids provide a uniform electric field in the active region to improve the accuracy of measurements and reduce ion migration time so as to virtually eliminate errors due to ion recombination. The parallel plate grids are fabricated with a minimal surface area to reduce the effects of contamination resulting from absorption of contaminating materials on the surface of the grids. Additionally, the ionization chamber wall is spaced a sufficient distance from the active region of the ionization chamber to minimize contamination effects.

  17. Search for lightly ionizing particles using CDMS-II data and fabrication of CDMS detectors with improved homogeneity in properties

    SciTech Connect

    Prasad, Kunj Bihari

    2013-12-01

    Fundamental particles are always observed to carry charges which are integral multiples of one-third charge of electron, e/3. While this is a well established experimental fact, the theoretical understanding for the charge quantization phenomenon is lacking. On the other hand, there exist numerous theoretical models that naturally allow for existence of particles with fractional electromagnetic charge. These particles, if existing, hint towards existence of physics beyond the standard model. Multiple high energy, optical, cosmological and astrophysical considerations restrict the allowable mass-charge parameter space for these fractional charges. Still, a huge unexplored region remains. The Cryogenic Dark Matter Search (CDMS-II), located at Soudan mines in northern Minnesota, employs germanium and silicon crystals to perform direct searches for a leading candidate to dark matter called Weakly Interacting Massive Particles (WIMPs). Alternately, the low detection threshold allows search for fractional electromagnetic-charged particles, or Lightly Ionizing Particles (LIPs), moving at relativistic speed. Background rejection is obtained by requiring that the magnitude and location of energy deposited in each detector be consistent with corresponding \\signatures" resulting from the passage of a fractionally charged particle. In this dissertation, the CDMS-II data is analyzed to search for LIPs, with an expected background of 0.078 0.078 events. No candidate events are observed, allowing exclusion of new parameter space for charges between e/6 and e/200.

  18. Smart smoke alarm

    DOEpatents

    Warmack, Robert J. Bruce; Wolf, Dennis A; Frank, Steven Shane

    2015-04-28

    Methods and apparatus for smoke detection are disclosed. In one embodiment, a smoke detector uses linear discriminant analysis (LDA) to determine whether observed conditions indicate that an alarm is warranted.

  19. Joint measurement of the atmospheric muon flux through the Puy de Dome volcano with plastic scintillators and Resistive Plate Chambers detectors

    SciTech Connect

    Ambrosino, F.; Anastasio, A.; Bross, A.; Bene, S.; Bonechi, L.; Carloganu, C.; Cimmino, L.; Combaret, Ch.; Durand, S.; Fehr, F.; Gailler, L.; Labazuy, Ph.; Laktineh, I.; Masone, V.; Miallier, D.; Mori, N.; Niess, V.; Pla-Dalmau, A.; Portal, A.; Rubinov, P.; Saracino, G.; Scarlini, E.; Strolin, P.; Vulpescu, B.

    2015-11-14

    The muographic imaging of volcanoes relies on the measured transmittance of the atmospheric muon flux through the target. An important bias affecting the result comes from background contamination mimicking a higher transmittance. The MU-RAY and TOMUVOL collaborations measured independently in 2013 the atmospheric muon flux transmitted through the Puy de Dôme volcano using their early prototype detectors, based on plastic scintillators and on Glass Resistive Plate Chambers, respectively. These detectors had three (MU-RAY) or four (TOMUVOL) detection layers of 1 m2 each, tens (MU-RAY) or hundreds (TOMUVOL) of nanosecond time resolution, a few millimeter position resolution, an energy threshold of few hundreds MeV, and no particle identification capabilities. The prototypes were deployed about 1.3 km away from the summit, where they measured, behind rock depths larger than 1000 m, remnant fluxes of 1.83±0.50(syst)±0.07(stat) m–2 d–1 deg–2 (MU-RAY) and 1.95±0.16(syst)±0.05(stat) m–2 d–1 deg–2 (TOMUVOL), that roughly correspond to the expected flux of high-energy atmospheric muons crossing 600 meters water equivalent (mwe) at 18° elevation. This implies that imaging depths larger than 500 mwe from 1 km away using such prototype detectors suffer from an overwhelming background. These measurements confirm that a new generation of detectors with higher momentum threshold, time-of-flight measurement, and/or particle identification is needed. As a result, the MU-RAY and TOMUVOL collaborations expect shortly to operate improved detectors, suitable for a robust muographic imaging of kilometer-scale volcanoes.

  20. Joint measurement of the atmospheric muon flux through the Puy de Dome volcano with plastic scintillators and Resistive Plate Chambers detectors

    DOE PAGES

    Ambrosino, F.; Anastasio, A.; Bross, A.; ...

    2015-11-14

    The muographic imaging of volcanoes relies on the measured transmittance of the atmospheric muon flux through the target. An important bias affecting the result comes from background contamination mimicking a higher transmittance. The MU-RAY and TOMUVOL collaborations measured independently in 2013 the atmospheric muon flux transmitted through the Puy de Dôme volcano using their early prototype detectors, based on plastic scintillators and on Glass Resistive Plate Chambers, respectively. These detectors had three (MU-RAY) or four (TOMUVOL) detection layers of 1 m2 each, tens (MU-RAY) or hundreds (TOMUVOL) of nanosecond time resolution, a few millimeter position resolution, an energy threshold ofmore » few hundreds MeV, and no particle identification capabilities. The prototypes were deployed about 1.3 km away from the summit, where they measured, behind rock depths larger than 1000 m, remnant fluxes of 1.83±0.50(syst)±0.07(stat) m–2 d–1 deg–2 (MU-RAY) and 1.95±0.16(syst)±0.05(stat) m–2 d–1 deg–2 (TOMUVOL), that roughly correspond to the expected flux of high-energy atmospheric muons crossing 600 meters water equivalent (mwe) at 18° elevation. This implies that imaging depths larger than 500 mwe from 1 km away using such prototype detectors suffer from an overwhelming background. These measurements confirm that a new generation of detectors with higher momentum threshold, time-of-flight measurement, and/or particle identification is needed. As a result, the MU-RAY and TOMUVOL collaborations expect shortly to operate improved detectors, suitable for a robust muographic imaging of kilometer-scale volcanoes.« less

  1. Evaluation of a novel helium ionization detector within the context of (low-)flow modulation comprehensive two-dimensional gas chromatography.

    PubMed

    Franchina, Flavio A; Maimone, Mariarosa; Sciarrone, Danilo; Purcaro, Giorgia; Tranchida, Peter Q; Mondello, Luigi

    2015-07-10

    The present research is focused on the use and evaluation of a novel helium ionization detector, defined as barrier discharge ionization detector (BID), within the context of (low-)flow modulation comprehensive two-dimensional gas chromatography (FM GC×GC). The performance of the BID device was compared to that of a flame ionization detector (FID), under similar FM GC×GC conditions. Following development and optimization of the FM GC×GC method, the BID was subjected to fine tuning in relation to acquisition frequency and discharge flow. Moreover, the BID performance was measured and compared to that of the FID, in terms of extra-column band broadening, sensitivity and dynamic range. The comparative study was carried out by using standard compounds belonging to different chemical classes, along with a sample of diesel fuel. Advantages and disadvantages of the BID system, also within the context of FM GC×GC, are critically discussed. In general, the BID system was characterized by a more limited dynamic range and increased sensitivity, compared to the FID. Additionally, BID and FID contribution to band broadening was found to be similar under the operational conditions applied. Particular attention was devoted to the behaviour of the FM GC×GC-BID system toward saturated and aromatic hydrocarbons, for a possible future use in the field of mineral-oil food contamination research.

  2. Calibration of impact ionization cosmic dust detectors: first tests to investigate how the dust density influences the signal

    NASA Astrophysics Data System (ADS)

    Jasmin Sterken, Veerle; Moragas-Klostermeyer, Georg; Hillier, Jon; Fielding, Lee; Lovett, Joseph; Armes, Steven; Fechler, Nina; Srama, Ralf; Bugiel, Sebastian; Hornung, Klaus

    2016-10-01

    Impact ionization experiments have been performed since more than 40 years for calibrating cosmic dust detectors. A linear Van de Graaff dust accelerator was used to accelerate the cosmic dust analogues of submicron to micron-size to speeds up to 80 km s^-1. Different materials have been used for calibration: iron, carbon, metal-coated minerals and most recently, minerals coated with conductive polymers. While different materials with different densities have been used for instrument calibration, a comparative analysis of dust impacts of equal material but different density is necessary: porous or aggregate-like particles are increasingly found to be present in the solar system: e.g. dust from comet 67P Churyumov-Gerasimenko [Fulle et al 2015], aggregate particles from the plumes of Enceladus [Gao et al 2016], and low-density interstellar dust [Westphal 2014 et al, Sterken et al 2015]. These recalibrations are relevant for measuring the size distributions of interplanetary and interstellar dust and thus mass budgets like the gas-to-dust mass ratio in the local interstellar cloud.We report about the calibrations that have been performed at the Heidelberg dust accelerator facility for investigating the influence of particle density on the impact ionization charge. We used the Cassini Cosmic Dust Analyzer for the target, and compared hollow versus compact silica particles in our study as a first attempt to investigate experimentally the influence of dust density on the signals obtained. Also, preliminary tests with carbon aerogel were performed, and (unsuccessful) attempts to accelerate silica aerogel. In this talk we explain the motivation of the study, the experiment set-up, the preparation of — and the materials used, the results and plans and recommendations for future tests.Fulle, M. et al 2015, The Astrophysical Journal Letters, Volume 802, Issue 1, article id. L12, 5 pp. (2015)Gao, P. et al 2016, Icarus, Volume 264, p. 227-238Westphal, A. et al 2014, Science

  3. The EXO-200 detector, part I: detector design and construction

    NASA Astrophysics Data System (ADS)

    Auger, M.; Auty, D. J.; Barbeau, P. S.; Bartoszek, L.; Baussan, E.; Beauchamp, E.; Benitez-Medina, C.; Breidenbach, M.; Chauhan, D.; Cleveland, B.; Conley, R.; Cook, J.; Cook, S.; Coppens, A.; Craddock, W.; Daniels, T.; Davis, C. G.; Davis, J.; deVoe, R.; Dobi, A.; Dolinski, M. J.; Dunford, M.; Fairbank, W., Jr.; Farine, J.; Fierlinger, P.; Franco, D.; Giroux, G.; Gornea, R.; Graham, K.; Gratta, G.; Hagemann, C.; Hall, C.; Hall, K.; Hargrove, C.; Herrin, S.; Hodgson, J.; Hughes, M.; Karelin, A.; Kaufman, L. J.; Kirk, J.; Kuchenkov, A.; Kumar, K. S.; Leonard, D. S.; Leonard, F.; LePort, F.; Mackay, D.; MacLellan, R.; Marino, M.; Merkle, K.; Mong, B.; Montero Díez, M.; Müller, A. R.; Neilson, R.; Odian, A.; O'Sullivan, K.; Ouellet, C.; Piepke, A.; Pocar, A.; Prescott, C. Y.; Pushkin, K.; Rivas, A.; Rollin, E.; Rowson, P. C.; Sabourov, A.; Sinclair, D.; Skarpaas, K.; Slutsky, S.; Stekhanov, V.; Strickland, V.; Swift, M.; Tosi, D.; Twelker, K.; Vuilleumier, J.-L.; Vuilleumier, J.-M.; Walton, T.; Weber, M.; Wichoski, U.; Wodin, J.; Wright, J. D.; Yang, L.; Yen, Y.-R.

    2012-05-01

    EXO-200 is an experiment designed to search for double beta decay of 136Xe with a single-phase, liquid xenon detector. It uses an active mass of 110 kg of xenon enriched to 80.6% in the isotope 136 in an ultra-low background time projection chamber capable of simultaneous detection of ionization and scintillation. This paper describes the EXO-200 detector with particular attention to the most innovative aspects of the design that revolve around the reduction of backgrounds, the efficient use of the expensive isotopically enriched xenon, and the optimization of the energy resolution in a relatively large volume.

  4. Three chamber negative ion source

    DOEpatents

    Leung, Ka-Ngo; Ehlers, Kenneth W.; Hiskes, John R.

    1985-01-01

    A negative ion vessel is divided into an excitation chamber, a negative ionization chamber and an extraction chamber by two magnetic filters. Input means introduces neutral molecules into a first chamber where a first electron discharge means vibrationally excites the molecules which migrate to a second chamber. In the second chamber a second electron discharge means ionizes the molecules, producing negative ions which are extracted into or by a third chamber. A first magnetic filter prevents high energy electrons from entering the negative ionization chamber from the excitation chamber. A second magnetic filter prevents high energy electrons from entering the extraction chamber from the negative ionizing chamber. An extraction grid at the end of the negative ion vessel attracts negative ions into the third chamber and accelerates them. Another grid, located adjacent to the extraction grid, carries a small positive voltage in order to inhibit positive ions from migrating into the extraction chamber and contour the plasma potential. Additional electrons can be suppressed from the output flux using ExB forces provided by magnetic field means and the extractor grid electric potential.

  5. NOTE: Hybrid plan verification for intensity-modulated radiation therapy (IMRT) using the 2D ionization chamber array I'mRT MatriXX—a feasibility study

    NASA Astrophysics Data System (ADS)

    Dobler, Barbara; Streck, Natalia; Klein, Elisabeth; Loeschel, Rainer; Haertl, Petra; Koelbl, Oliver

    2010-01-01

    The 2D ionization chamber array I'mRT MatriXX (IBA, Schwarzenbruck, Germany) has been developed for absolute 2D dosimetry and verification of intensity-modulated radiation therapy (IMRT) for perpendicular beam incidence. The aim of this study is to evaluate the applicability of I'mRT MatriXX for oblique beam incidence and hybrid plan verification of IMRT with original gantry angles. For the assessment of angular dependence, open fields with gantry angles in steps of 10° were calculated on a CT scan of I'mRT MatriXX. For hybrid plan verification, 17 clinical IMRT plans and one rotational plan were used. Calculations were performed with pencil beam (PB), collapsed cone (CC) and Monte Carlo (MC) methods, which had been previously validated. Measurements were conducted on an Elekta SynergyS linear accelerator. To assess the potential and limitations of the system, gamma evaluation was performed with different dose tolerances and distances to agreement. Hybrid plan verification passed the gamma test with 4% dose tolerance and 3 mm distance to agreement in all cases, in 82-88% of the cases for tolerances of 3%/3 mm, and in 59-76% of the cases if 3%/2 mm were used. Separate evaluation of the low dose and high dose regions showed that I'mRT MatriXX can be used for hybrid plan verification of IMRT plans within 3% dose tolerance and 3 mm distance to agreement with a relaxed dose tolerance of 4% in the low dose region outside the multileaf collimator (MLC).

  6. Chemistry of α-pinene and naphthalene oxidation products generated in a Potential Aerosol Mass (PAM) chamber as measured by acetate chemical ionization mass spectrometry

    DOE PAGES

    Chhabra, P. S.; Lambe, A. T.; Canagaratna, M. R.; ...

    2014-07-01

    Recent developments in high resolution, time-of-flight chemical ionization mass spectrometry (HR-ToF-CIMS) have made possible the direct detection of atmospheric organic compounds in real-time with high sensitivity and with little or no fragmentation, including low volatility, highly oxygenated organic vapors that are precursors to secondary organic aerosol formation. Here, for the first time, we examine gas-phase O3 and OH oxidation products of α-pinene and naphthalene formed in the PAM flow reactor with an HR-ToF-CIMS using acetate reagent ion chemistry. Integrated OH exposures ranged from 1.2 × 1011 to 9.7 × 1011 molec cm−3 s, corresponding to approximately 1.0 to 7.5 daysmore » of equivalent atmospheric oxidation. Measured gas-phase organic acids are similar to those previously observed in environmental chamber studies. For both precursors, we find that acetate-CIMS spectra capture both functionalization (oxygen addition) and fragmentation (carbon loss) as a function of OH exposure. The level of fragmentation is observed to increase with increased oxidation. We present a method that estimates vapor pressures of organic molecules using the measured O/C ratio, H/C ratio, and carbon number for each compound detected by the CIMS. The predicted condensed-phase SOA average acid yields and O/C and H/C ratios agree within uncertainties with previous AMS measurements and ambient CIMS results. While acetate reagent ion chemistry is used to selectively measure organic acids, in principle this method can be applied to additional reagent ion chemistries depending on the application.« less

  7. Readout channel with majority logic timestamp and digital peak detector for Muon Chambers of the CBM experiment

    NASA Astrophysics Data System (ADS)

    Atkin, E.; Bulbakov, I.; Ivanov, P.; Ivanov, V.; Malankin, E.; Normanov, D.; Sagdiev, I.; Samsonov, V.; Shumikhin, V.; Shumkin, O.; Vinogradov, S.; Voronin, A.

    2016-12-01

    A prototype readout channel was manufactured in UMC CMOS 180 nm for the purpose of the CBM experiment at the FAIR accelerator. The channel includes a preamplifier with fast and slow CR-RC shapers, discriminator with a differential threshold setup circuit, a 6-bit SAR ADC (DNL = 0.70, INL = 0.45), digital peak detector and block of the time stamp registration. The control data, clock and output data are supplied through SLVS transmitter and receiver. The slow and fast channels have 1000 el and 1500 el ENC accordingly at a 50 pF detector capacitance. Power consumption is 10 mW/channel.

  8. Selective determination of tobacco-specific nitrosamines in mainstream cigarette smoke by GC coupled to positive chemical ionization triple quadrupole MS.

    PubMed

    Wu, Da; Lu, Yifeng; Lin, Huaqing; Zhou, Wanhong; Gu, Wenbo

    2013-08-01

    A rapid method for the selective determination of four kinds of tobacco-specific nitrosamines, N-nitrosonornicotine, N-nitrosoanatabine, N-nitrosoanabasine and 4-(methylnitrosamino)-1-(3-pyridyl)-1-butanone, in mainstream cigarette smoke was developed by GC coupled to positive chemical ionization triple-quadrupole MS. After mainstream cigarette smoke was collected on a cambridge filter pad, the particulate matter was extracted with 0.1 M HCL aqueous solution, cleaned by positive cation-exchange solid extraction, and finally injected into GC-MS/MS using isotopically labeled analogues as internal standards. Excellent linearity was obtained over the concentration range of 0.5-200.0 ng mL(-1) for all tobacco-specific nitrosamines with values for correlation coefficient between 0.9996-0.9999. Limits of detection of each tobacco specific nitrosamine varied from 0.023-0.028 ng cig(-1), and lower limits of quantification varied from 0.077-0.093 ng cig(-1). The recovery of each tobacco specific nitrosamine was from 90.0-109.0%. The relative standard deviations of the intra-day and inter-day precisions were 3.1-5.8 and 3.9-6.6, respectively. This method was applied to reference and domestic cigarettes. The result showed that the method was consistent with traditional methods and can be used as an effective approach for the routine analysis of tobacco-specific nitrosamines.

  9. UHPLC separation with MS analysis for eight carbonyl compounds in mainstream tobacco smoke.

    PubMed

    Miller, John H; Gardner, William P; Gonzalez, Ricardo R

    2010-01-01

    A method to quantify eight carbonyl compounds in mainstream cigarette smoke is presented using ultra-high pressure liquid chromatography (UHPLC). The combination of UHPLC and mass spectrometry (UHPLC-MS) dramatically reduces analysis times as compared to the current in-house high-performance liquid chromatography (HPLC)-UV method. In addition, improved detector selectivity and peak resolution are observed. Sample analysis times are reduced from 47 min with HPLC-UV to less than 5 min using this improved method. Atmospheric pressure chemical ionization, atmospheric pressure photo ionization, and electrospray ionization are directly compared to evaluate ionization potential and linear response range for the carbonyl 2,4-dinitrophenylhydrazine derivatives. Smoke extracts from three standard smoking protocols are analyzed by both UHPLC-MS and HPLC-UV for method comparison purposes.

  10. DETECTORS AND EXPERIMENTAL METHODS: Design and test of a Multi-gap Resistive Plate Chamber with Long readout-strip (LMRPC)

    NASA Astrophysics Data System (ADS)

    Sun, Yong-Jie; Li, Cheng; Zhou, Yi; Shao, Ming; Zhao, Yan-E.; Chen, Hong-Fang

    2009-02-01

    A new kind (two end readout) of Multi-gap Resistive Plate Chamber with long readout-strip (LMRPC) is developed to be used at the large-area Muon Telescope Detector (MTD) at mid-rapidity at RHIC/STAR experiment for Time-of-Flight (TOF) measurement. The LMRPC has an active area of 87 cm × 17 cm, 10 gas gaps of 250 μm arranged in 2 stacks, with readout strips of 2.5 cm wide and 90 cm long. The considerations in LMRPC design related to its performance are discussed in this paper. The cosmic ray test results of a prototype LMRPC show a detection efficiency >95% and the time resolution ~70 ps.

  11. 10 CFR 32.14 - Certain items containing byproduct material; requirements for license to apply or initially...

    Code of Federal Regulations, 2014 CFR

    2014-01-01

    ... electron tubes and ionization chamber smoke detectors and timepieces containing promethium-147 or tritium... dials containing tritium or promethium-147, and its container with the identification of...

  12. 10 CFR 32.14 - Certain items containing byproduct material; requirements for license to apply or initially...

    Code of Federal Regulations, 2013 CFR

    2013-01-01

    ... electron tubes and ionization chamber smoke detectors and timepieces containing promethium-147 or tritium... dials containing tritium or promethium-147, and its container with the identification of...

  13. Static diffusion cloud chambers

    NASA Technical Reports Server (NTRS)

    Ayers, G.

    1981-01-01

    The chamber geometry and optical arrangement are described. The supersaturation range is given and consists of readings taken at five fixed points: 0.25%, 0.5%, 0.75%, 1.0%, and 1.25%. The detection system is described including light source, cameras, and photocell detectors. The temperature control and the calibration of the chamber are discussed.

  14. Applications of gaseous particle detectors in physics and medicine

    NASA Astrophysics Data System (ADS)

    Sauli, Fabio

    1995-08-01

    The multi-wire proportional chamber, introduced in 1967 by Georges Charpak (recipient of the 1992 Nobel prize for physics) allows to achieve high-rate, fully electronics detection and localization of ionizing radiation. The myriad of devices inspired by this initial work generated a revolution in the conception of detectors for elementary particle physics experiments; examples are the time projection chamber, the drift chamber, the micro-strip gas chamber. After a brief introduction on the basic operating principles of the device, I will describe several examples of application of advanced gas detectors in medicine and biology and analyze the operating characteristics that make the new devices attractive when confronted with classic detectors.

  15. SU-E-T-35: An Investigation of the Accuracy of Cervical IMRT Dose Distribution Using 2D/3D Ionization Chamber Arrays System and Monte Carlo Simulation

    SciTech Connect

    Zhang, Y; Yang, J; Liu, H; Liu, D

    2014-06-01

    Purpose: The purpose of this work is to compare the verification results of three solutions (2D/3D ionization chamber arrays measurement and Monte Carlo simulation), the results will help make a clinical decision as how to do our cervical IMRT verification. Methods: Seven cervical cases were planned with Pinnacle 8.0m to meet the clinical acceptance criteria. The plans were recalculated in the Matrixx and Delta4 phantom with the accurate plans parameters. The plans were also recalculated by Monte Carlo using leaf sequences and MUs for individual plans of every patient, Matrixx and Delta4 phantom. All plans of Matrixx and Delta4 phantom were delivered and measured. The dose distribution of iso slice, dose profiles, gamma maps of every beam were used to evaluate the agreement. Dose-volume histograms were also compared. Results: The dose distribution of iso slice and dose profiles from Pinnacle calculation were in agreement with the Monte Carlo simulation, Matrixx and Delta4 measurement. A 95.2%/91.3% gamma pass ratio was obtained between the Matrixx/Delta4 measurement and Pinnacle distributions within 3mm/3% gamma criteria. A 96.4%/95.6% gamma pass ratio was obtained between the Matrixx/Delta4 measurement and Monte Carlo simulation within 2mm/2% gamma criteria, almost 100% gamma pass ratio within 3mm/3% gamma criteria. The DVH plot have slightly differences between Pinnacle and Delta4 measurement as well as Pinnacle and Monte Carlo simulation, but have excellent agreement between Delta4 measurement and Monte Carlo simulation. Conclusion: It was shown that Matrixx/Delta4 and Monte Carlo simulation can be used very efficiently to verify cervical IMRT delivery. In terms of Gamma value the pass ratio of Matrixx was little higher, however, Delta4 showed more problem fields. The primary advantage of Delta4 is the fact it can measure true 3D dosimetry while Monte Carlo can simulate in patients CT images but not in phantom.

  16. Cost effective, dual-purpose machine vision-based detectors for (1) smoke and flame detection, and (2) engine overhead/burn-through and flame detection

    NASA Astrophysics Data System (ADS)

    Goedeke, A. D.; Drda, B.; Fitzpatrick, C.; Healey, G.

    1995-03-01

    Three major accomplishments resulted from this effort: (1) It was determined that the Machine Vision Fire Detection System (MVFDS) could be augmented with a near-infrared imaging FPA/CCD, 4-8 IR transmitting optic waveguides, and modified algorithms to measure small changes in temperature of small areas on the engine casing and therefore determine overheat condition and/or possible burn-through condition. The IR-MVFDS would also detect small burn-through torch-like fires as well as nacelle fires associated with munitions and leaking/broken fuel and hydraulic lines. (2) Tests verified the ability of the visible MVFDS flame detector to simultaneously detect smoke and measure its transmissivity. This dual function, in addition to possibly as many as 16 or more separate fields-of-view using small fiber optic cables, makes the Smoke/Flame-MVFDS cost effective in aircraft and in commercial applications. (3) The potential benefits and cost savings of both the IR-MVFDS and Smoke/Flame-MVFDS were demonstrated. The technical basis was proven feasible with minimum risk for development of both the IR-MVFDS and the Flame/Smoke-MVFDS in Phase II.

  17. Application of gas chromatography/flame ionization detector-based metabolite fingerprinting for authentication of Asian palm civet coffee (Kopi Luwak).

    PubMed

    Jumhawan, Udi; Putri, Sastia Prama; Yusianto; Bamba, Takeshi; Fukusaki, Eiichiro

    2015-11-01

    Development of authenticity screening for Asian palm civet coffee, the world-renowned priciest coffee, was previously reported using metabolite profiling through gas chromatography/mass spectrometry (GC/MS). However, a major drawback of this approach is the high cost of the instrument and maintenance. Therefore, an alternative method is needed for quality and authenticity evaluation of civet coffee. A rapid, reliable and cost-effective analysis employing a universal detector, GC coupled with flame ionization detector (FID), and metabolite fingerprinting has been established for discrimination analysis of 37 commercial and non-commercial coffee beans extracts. gas chromatography/flame ionization detector (GC/FID) provided higher sensitivity over a similar range of detected compounds than GC/MS. In combination with multivariate analysis, GC/FID could successfully reproduce quality prediction from GC/MS for differentiation of commercial civet coffee, regular coffee and coffee blend with 50 wt % civet coffee content without prior metabolite details. Our study demonstrated that GC/FID-based metabolite fingerprinting can be effectively actualized as an alternative method for coffee authenticity screening in industries.

  18. SU-E-T-552: Monte Carlo Calculation of Correction Factors for a Free-Air Ionization Chamber in Support of a National Air-Kerma Standard for Electronic Brachytherapy

    SciTech Connect

    Mille, M; Bergstrom, P

    2015-06-15

    Purpose: To use Monte Carlo radiation transport methods to calculate correction factors for a free-air ionization chamber in support of a national air-kerma standard for low-energy, miniature x-ray sources used for electronic brachytherapy (eBx). Methods: The NIST is establishing a calibration service for well-type ionization chambers used to characterize the strength of eBx sources prior to clinical use. The calibration approach involves establishing the well-chamber’s response to an eBx source whose air-kerma rate at a 50 cm distance is determined through a primary measurement performed using the Lamperti free-air ionization chamber. However, the free-air chamber measurements of charge or current can only be related to the reference air-kerma standard after applying several corrections, some of which are best determined via Monte Carlo simulation. To this end, a detailed geometric model of the Lamperti chamber was developed in the EGSnrc code based on the engineering drawings of the instrument. The egs-fac user code in EGSnrc was then used to calculate energy-dependent correction factors which account for missing or undesired ionization arising from effects such as: (1) attenuation and scatter of the x-rays in air; (2) primary electrons escaping the charge collection region; (3) lack of charged particle equilibrium; (4) atomic fluorescence and bremsstrahlung radiation. Results: Energy-dependent correction factors were calculated assuming a monoenergetic point source with the photon energy ranging from 2 keV to 60 keV in 2 keV increments. Sufficient photon histories were simulated so that the Monte Carlo statistical uncertainty of the correction factors was less than 0.01%. The correction factors for a specific eBx source will be determined by integrating these tabulated results over its measured x-ray spectrum. Conclusion: The correction factors calculated in this work are important for establishing a national standard for eBx which will help ensure that dose

  19. Quantification of Triacylglycerol Molecular Species in Edible Fats and Oils by Gas Chromatography-Flame Ionization Detector Using Correction Factors.

    PubMed

    Yoshinaga, Kazuaki; Obi, Junji; Nagai, Toshiharu; Iioka, Hiroyuki; Yoshida, Akihiko; Beppu, Fumiaki; Gotoh, Naohiro

    2017-03-01

    In the present study, the resolution parameters and correction factors (CFs) of triacylglycerol (TAG) standards were estimated by gas chromatography-flame ionization detector (GC-FID) to achieve the precise quantification of the TAG composition in edible fats and oils. Forty seven TAG standards comprising capric acid, lauric acid, myristic acid, pentadecanoic acid, palmitic acid, palmitoleic acid, stearic acid, oleic acid, linoleic acid, and/or linolenic acid were analyzed, and the CFs of these TAGs were obtained against tripentadecanoyl glycerol as the internal standard. The capillary column was Ultra ALLOY(+)-65 (30 m × 0.25 mm i.d., 0.10 μm thickness) and the column temperature was programmed to rise from 250°C to 360°C at 4°C/min and then hold for 25 min. The limit of detection (LOD) and limit of quantification (LOQ) values of the TAG standards were > 0.10 mg and > 0.32 mg per 100 mg fat and oil, respectively, except for LnLnLn, and the LOD and LOQ values of LnLnLn were 0.55 mg and 1.84 mg per 100 mg fat and oil, respectively. The CFs of TAG standards decreased with increasing total acyl carbon number and degree of desaturation of TAG molecules. Also, there were no remarkable differences in the CFs between TAG positional isomers such as 1-palmitoyl-2-oleoyl-3-stearoyl-rac-glycerol, 1-stearoyl-2-palmitoyl-3-oleoyl-rac-glycerol, and 1-palmitoyl-2-stearoyl-3-oleoyl-rac-glycerol, which cannot be separated by GC-FID. Furthermore, this method was able to predict the CFs of heterogeneous (AAB- and ABC-type) TAGs from the CFs of homogenous (AAA-, BBB-, and CCC-type) TAGs. In addition, the TAG composition in cocoa butter, palm oil, and canola oil was determined using CFs, and the results were found to be in good agreement with those reported in the literature. Therefore, the GC-FID method using CFs can be successfully used for the quantification of TAG molecular species in natural fats and oils.

  20. Comparison of a gas chromatography-optical fibre (GC-OF) detector with a gas chromatography-flame ionization detector (GC-FID) for determination of alcoholic compounds in industrial atmospheres.

    PubMed

    Silva, Lurdes I B; Rocha-Santos, Teresa A P; Duarte, A C

    2008-07-15

    An analytical methodology based on an optical fibre detector coupled to gas chromatograph has been developed for the speciation of some volatile alcoholic compounds. This methodology combines the separation capability of gas chromatography with an optical fibre detector made of an optical fibre sensitized with a thin polymeric film of poly[methyl(3,3,3-trifluoropropyl)siloxane] (PMTFPS). The response of the detector has been characterized at 650 nm for nine different alcohols (allyl alcohol, n-propyl alcohol, sec-butyl alcohol, isobutyl alcohol, n-butyl alcohol, isoamyl alcohol, methyl isobutyl carbinol, cyclohexanol and diacetone alcohol). An alternative method based on gas chromatography-flame ionization detector (GC-FID) was also used in order to evaluated the performance and compare the analytical results with the proposed method. The time of analysis, the analytical error and the analytical performance were similar for both methods. However, the analytical apparatus based on the GC-OF detector is much less expensive than the GC-FID and show high accuracy and suitability for actual monitoring on indoor atmospheres.

  1. Online monitor detector for the protontherapy beam at the INFN Laboratori Nazionali del Sud-Catania

    NASA Astrophysics Data System (ADS)

    Givehchi, N.; Marchetto, F.; Boriano, A.; Attili, A.; Bourhaleb, F.; Cirio, R.; Cirrone, G. A. P.; Cuttone, G.; Di Rosa, F.; Donetti, M.; Garella, M. A.; Giordanengo, S.; Iliescu, S.; La Rosa, A.; Lojacono, P. A.; Nicotra, P.; Peroni, C.; Pecka, A.; Pitta, G.; Raffaele, L.; Russo, G.; Sabini, M. G.; Valastro, L. M.

    2007-03-01

    A detector to monitor online the protontherapy beam at the Istituto Nazionale di Fisica Nucleare (INFN), Laboratori Nazionali del Sud (LNS) has been built and characterized. The detector is made of two ionization chambers: each chamber has the anode splitted into 256 0.5 mm strips with vertical and horizontal orientation. The chambers are part of the beam line: signals can be processed online at a speed up to 100 Hz and results are promptly available. Thus the beam geometry can be controlled continuously during patient treatment, and in case of deviation from the required conditions, the treatment can be directly concluded.

  2. Smoke detection

    DOEpatents

    Warmack, Robert J. Bruce; Wolf, Dennis A.; Frank, Steven Shane

    2015-10-27

    Various apparatus and methods for smoke detection are disclosed. In one embodiment, a method of training a classifier for a smoke detector comprises inputting sensor data from a plurality of tests into a processor. The sensor data is processed to generate derived signal data corresponding to the test data for respective tests. The derived signal data is assigned into categories comprising at least one fire group and at least one non-fire group. Linear discriminant analysis (LDA) training is performed by the processor. The derived signal data and the assigned categories for the derived signal data are inputs to the LDA training. The output of the LDA training is stored in a computer readable medium, such as in a smoke detector that uses LDA to determine, based on the training, whether present conditions indicate the existence of a fire.

  3. Smoke detection

    DOEpatents

    Warmack, Robert J. Bruce; Wolf, Dennis A.; Frank, Steven Shane

    2016-09-06

    Various apparatus and methods for smoke detection are disclosed. In one embodiment, a method of training a classifier for a smoke detector comprises inputting sensor data from a plurality of tests into a processor. The sensor data is processed to generate derived signal data corresponding to the test data for respective tests. The derived signal data is assigned into categories comprising at least one fire group and at least one non-fire group. Linear discriminant analysis (LDA) training is performed by the processor. The derived signal data and the assigned categories for the derived signal data are inputs to the LDA training. The output of the LDA training is stored in a computer readable medium, such as in a smoke detector that uses LDA to determine, based on the training, whether present conditions indicate the existence of a fire.

  4. Development of a large area InGaAs APD receiver based on an impact ionization engineered detector for free-space lasercomm applications

    NASA Astrophysics Data System (ADS)

    Burris, H. R.; Ferraro, M. S.; Freeman, W. T.; Moore, C. I.; Murphy, J. L.; Rabinovich, W. S.; Smith, W. R.; Summers, L. L.; Thomas, L. M.; Vilcheck, M. J.; Clark, W. R.; Waters, W. D.

    2012-06-01

    The U.S. Naval Research Laboratory (NRL) is developing a small size, weight and power (SWaP) free space lasercomm terminal for small unmanned airborne platforms. The terminal is based on a small gimbal developed by CloudCap Technology. A receiver with a large field of view and with sensitivity sufficient to meet the program range goals is required for this terminal. An InGaAs Avalanche Photodiode (APD) with internal structures engineered to reduce excess noise and keff in high gain applications was selected as the detector. The detector is a 350 micron diameter impact ionization engineered (I2E) APD developed by Optogration, Inc. Results of development and characterization of the receiver will be presented.

  5. Flow chamber

    SciTech Connect

    Morozov, Victor

    2011-01-18

    A flow chamber having a vacuum chamber and a specimen chamber. The specimen chamber may have an opening through which a fluid may be introduced and an opening through which the fluid may exit. The vacuum chamber may have an opening through which contents of the vacuum chamber may be evacuated. A portion of the flow chamber may be flexible, and a vacuum may be used to hold the components of the flow chamber together.

  6. Practical method for determination of air kerma by use of an ionization chamber toward construction of a secondary X-ray field to be used in clinical examination rooms.

    PubMed

    Maehata, Itsumi; Hayashi, Hiroaki; Kimoto, Natsumi; Takegami, Kazuki; Okino, Hiroki; Kanazawa, Yuki; Tominaga, Masahide

    2016-07-01

    We propose a new practical method for the construction of an accurate secondary X-ray field using medical diagnostic X-ray equipment. For accurate measurement of the air kerma of an X-ray field, it is important to reduce and evaluate the contamination rate of scattered X-rays. To determine the rate quantitatively, we performed the following studies. First, we developed a shield box in which an ionization chamber could be set at an inner of the box to prevent detection of the X-rays scattered from the air. In addition, we made collimator plates which were placed near the X-ray source for estimation of the contamination rate by scattered X-rays from the movable diaphragm which is a component of the X-ray equipment. Then, we measured the exposure dose while changing the collimator plates, which had diameters of 25-90 mm(ϕ). The ideal value of the exposure dose was derived mathematically by extrapolation to 0 mm(ϕ). Tube voltages ranged from 40 to 130 kV. Under these irradiation conditions, we analyzed the contamination rate by the scattered X-rays. We found that the contamination rates were less than 1.7 and 2.3 %, caused by air and the movable diaphragm, respectively. The extrapolated value of the exposure dose has been determined to have an uncertainty of 0.7 %. The ionization chamber used in this study was calibrated with an accuracy of 5 %. Using this kind of ionization chamber, we can construct a secondary X-ray field with an uncertainty of 5 %.

  7. A case-cohort study of lung cancer, ionizing radiation, and tobacco smoking among males at the Hanford Site

    SciTech Connect

    Petersen, G.R.; Gilbert, E.S.; Buchanan, J.A.; Stevens, R.G. )

    1990-01-01

    Results of several epidemiological studies of workers exposed occupationally to low levels of radiation have been reported but have not included data on smoking. The authors conducted a case-cohort study of male workers at the Hanford Site with an objective of investigating the association between lung-cancer risk and occupational radiation exposure with appropriate adjustment for tobacco use. Eighty-six lung-cancer deaths for the period 1965-1980 and a stratified random sample of 445 subcohort members were included in the study. Tobacco-use data were obtained from medical records collected over each subject's period of employment. Data from this study were analyzed using methods that took into account both the case-cohort design and changes over time in the quality of the tobacco-use data collected. Tobacco use was not strongly related to the level of radiation exposure, and adjustment for tobacco use did not greatly modify results of analyses assessing the association between lung-cancer risk and cumulative dose equivalent. With or without adjustment for tobacco use, the estimated risks per unit of cumulative dose equivalent were negative, but the 95% confidence intervals were wide and included values several times those estimated from populations with high levels of irradiation.

  8. Calculation of reactivities using ionization chamber currents with different sets of kinetic parameters for reduced scram system efficiency in the VVER-1000 of the third unit of the Kalinin nuclear power plant at the stage of physical start-up

    SciTech Connect

    Zizin, M. N.; Zizina, S. N.; Kryakvin, L. V.; Pitilimov, V. A.; Tereshonok, V. A.

    2011-12-15

    The effectiveness of the VVER-1000 reactor scram system is analyzed using ionization chamber currents with different sets of kinetic parameters with allowance for the isotopic composition in the calculation of these parameters. The most 'correct, aesthetically acceptable' results are obtained using the eight-group constants of the ROSFOND (BNAB-RF) library. The difference between the maximum and minimum values of the scram system effectiveness calculated with different sets of kinetic parameters slightly exceeds 2{beta}. The problems of introducing corrections due to spatial effects are not considered in this study.

  9. Invited review article: physics and Monte Carlo techniques as relevant to cryogenic, phonon, and ionization readout of Cryogenic Dark Matter Search radiation detectors.

    PubMed

    Leman, Steven W

    2012-09-01

    This review discusses detector physics and Monte Carlo techniques for cryogenic, radiation detectors that utilize combined phonon and ionization readout. A general review of cryogenic phonon and charge transport is provided along with specific details of the Cryogenic Dark Matter Search detector instrumentation. In particular, this review covers quasidiffusive phonon transport, which includes phonon focusing, anharmonic decay, and isotope scattering. The interaction of phonons in the detector surface is discussed along with the downconversion of phonons in superconducting films. The charge transport physics include a mass tensor which results from the crystal band structure and is modeled with a Herring-Vogt transformation. Charge scattering processes involve the creation of Neganov-Luke phonons. Transition-edge-sensor (TES) simulations include a full electric circuit description and all thermal processes including Joule heating, cooling to the substrate, and thermal diffusion within the TES, the latter of which is necessary to model normal-superconducting phase separation. Relevant numerical constants are provided for these physical processes in germanium, silicon, aluminum, and tungsten. Random number sampling methods including inverse cumulative distribution function (CDF) and rejection techniques are reviewed. To improve the efficiency of charge transport modeling, an additional second order inverse CDF method is developed here along with an efficient barycentric coordinate sampling method of electric fields. Results are provided in a manner that is convenient for use in Monte Carlo and references are provided for validation of these models.

  10. Invited Review Article: Physics and Monte Carlo techniques as relevant to cryogenic, phonon, and ionization readout of Cryogenic Dark Matter Search radiation detectors

    SciTech Connect

    Leman, Steven W.

    2012-09-15

    This review discusses detector physics and Monte Carlo techniques for cryogenic, radiation detectors that utilize combined phonon and ionization readout. A general review of cryogenic phonon and charge transport is provided along with specific details of the Cryogenic Dark Matter Search detector instrumentation. In particular, this review covers quasidiffusive phonon transport, which includes phonon focusing, anharmonic decay, and isotope scattering. The interaction of phonons in the detector surface is discussed along with the downconversion of phonons in superconducting films. The charge transport physics include a mass tensor which results from the crystal band structure and is modeled with a Herring-Vogt transformation. Charge scattering processes involve the creation of Neganov-Luke phonons. Transition-edge-sensor (TES) simulations include a full electric circuit description and all thermal processes including Joule heating, cooling to the substrate, and thermal diffusion within the TES, the latter of which is necessary to model normal-superconducting phase separation. Relevant numerical constants are provided for these physical processes in germanium, silicon, aluminum, and tungsten. Random number sampling methods including inverse cumulative distribution function (CDF) and rejection techniques are reviewed. To improve the efficiency of charge transport modeling, an additional second order inverse CDF method is developed here along with an efficient barycentric coordinate sampling method of electric fields. Results are provided in a manner that is convenient for use in Monte Carlo and references are provided for validation of these models.

  11. The emulsion chamber technology experiment

    NASA Technical Reports Server (NTRS)

    Gregory, John C.

    1992-01-01

    Photographic emulsion has the unique property of recording tracks of ionizing particles with a spatial precision of 1 micron, while also being capable of deployment over detector areas of square meters or 10's of square meters. Detectors are passive, their cost to fly in Space is a fraction of that of instruments of similar collecting. A major problem in their continued use has been the labor intensiveness of data retrieval by traditional microscope methods. Two factors changing the acceptability of emulsion technology in space are the astronomical costs of flying large electronic instruments such as ionization calorimeters in Space, and the power and low cost of computers, a small revolution in the laboratory microscope data-taking. Our group at UAH made measurements of the high energy composition and spectra of cosmic rays. The Marshall group has also specialized in space radiation dosimetry. Ionization calorimeters, using alternating layers of lead and photographic emulsion, to measure particle energies up to 10(exp 15) eV were developed. Ten balloon flights were performed with them. No such calorimeters have ever flown in orbit. In the ECT program, a small emulsion chamber was developed and will be flown on the Shuttle mission OAST-2 to resolve the principal technological questions concerning space exposures. These include assessments of: (1) pre-flight and orbital exposure to background radiation, including both self-shielding and secondary particle generation; the practical limit to exposure time in space can then be determined; (2) dynamics of stack to optimize design for launch and weightlessness; and (3) thermal and vacuum constraints on emulsion performance. All these effects are cumulative and affect our ability to perform scientific measurements but cannot be adequately predicted by available methods.

  12. Smoke detection in low-G fires

    NASA Technical Reports Server (NTRS)

    Urban, David L.; Griffin, Devon W.; Gard, Melissa Y.; Hoy, Michael

    1995-01-01

    Fires in spacecraft are considered a credible risk. To respond to this risk, NASA flew fire detectors on Skylab and the Space Shuttle (STS) and included them in the design for International Space Station Alpha (ISSA). In previous missions (Mercury, Gemini and Apollo), the crew quarters were so cramped that it was not considered credible that the astronauts could fail to observe a fire. The Skylab nodule included approximately 20 UV fire detectors. The space shuttle has 9 ionization detectors in the mid deck and flight deck and Spacelab has six additional ionization detectors. The planned detectors for ISSA are laser-diode, forward-scattering, smoke or particulate detectors. Current plans for the ISSA call for two detectors in the open area of the module and detectors in racks that have both cooling air flow and electrical power. Due to the complete absence of data concerning the nature of particulate and radiant emission from low-g fires, all three of these detector systems were designed based upon 1-g test data. As planned mission durations and complexity increase and the volume of spacecraft increases, the need for and importance of effective, crew independent, fire detection grows significantly. This requires more knowledge concerning low-gravity fires and how they might be detected. To date, no combustion-generated particulate samples have been collected for well-developed microgravity flames. All of the extant data come from drop tower tests and therefore only correspond to the early stages of a fire. The fuel sources were restricted to laminar gas-jet diffusion flames and rapidly overheated wire insulation. These gas-jet drop tower tests indicate, through thermophoretic sampling, that soot primaries and aggregates (groups of primary particles) in micro-g may be significantly larger than those in normal-g (ng). This raises new scientific questions about soot processes as well as practical issues for particulate detection/alarm threshold levels used in on

  13. Wide-range radioactive-gas-concentration detector

    DOEpatents

    Anderson, D.F.

    1981-11-16

    A wide-range radioactive-gas-concentration detector and monitor capable of measuring radioactive-gas concentrations over a range of eight orders of magnitude is described. The device is designed to have an ionization chamber sufficiently small to give a fast response time for measuring radioactive gases but sufficiently large to provide accurate readings at low concentration levels. Closely spaced parallel-plate grids provide a uniform electric field in the active region to improve the accuracy of measurements and reduce ion migration time so as to virtually eliminate errors due to ion recombination. The parallel-plate grids are fabricated with a minimal surface area to reduce the effects of contamination resulting from absorption of contaminating materials on the surface of the grids. Additionally, the ionization-chamber wall is spaced a sufficient distance from the active region of the ionization chamber to minimize contamination effects.

  14. Ethanol Concentration in 56 Refillable Electronic Cigarettes Liquid Formulations Determined by Headspace Gas Chromatography with Flame Ionization Detector (HS-GC-FID).

    PubMed

    Poklis, Justin L; Wolf, Carl E; Peace, Michelle R

    2017-03-23

    Personal battery powered vaporizers or electronic cigarettes were developed as an alternative to traditional cigarettes. The modern electronic cigarettes were patented in 2004 by Hon Lik in China. In May 2016, the US Food and Drug Administration (FDA) imposed regulatory statutes on e-cigarettes and their liquid formulations (e-liquids); prior to that, they were unregulated. E-liquids are typically composed of propylene glycol and/or glycerin, flavoring component(s), and active ingredient(s), such as nicotine. Fifty-six commercially available e-liquids, purchased from various sources, contained a variety of flavors and active ingredients. A headspace gas chromatography with flame ionization detector HS-GC-FID method was used to analyze these e-liquids for volatiles content. Only one of the e-liquids listed ethanol as a component. A headspace gas chromatography with flame ionization detector HS-GC-FID method was used to analyze the e-liquids. The chromatographic separation of volatiles was performed on a Restek BAC-1 column. A linear calibration was generated for ethanol with limits of detection and quantification (LOD/LOQ) of 0.05 mg/mL. Ethanol concentrations in the fifty-six e-liquids ranged from none detected to 206 mg/mL. The ethanol determined in these products may have been used in flavorants or a solvent; the reason for inclusion cannot be fully ascertained. The implications of vaporizing ethanol as an e-liquid component are unknown.

  15. Short-term chamber exposure to low doses of two kinds of wood smoke does not induce systemic inflammation, coagulation or oxidative stress in healthy humans

    PubMed Central

    Sallsten, Gerd; Almerud, Pernilla; Basu, Samar; Barregard, Lars

    2013-01-01

    Introduction: Air pollution increases the risk of cardiovascular diseases. A proposed mechanism is that local airway inflammation leads to systemic inflammation, affecting coagulation and the long-term risk of atherosclerosis. One major source of air pollution is wood burning. Here we investigate whether exposure to two kinds of wood smoke, previously shown to cause airway effects, affects biomarkers of systemic inflammation, coagulation and lipid peroxidation. Methods: Thirteen healthy adults were exposed to filtered air followed by two sessions of wood smoke for three hours, one week apart. One session used smoke from the start-up phase of the wood-burning cycle, and the other smoke from the burn-out phase. Mean particle mass concentrations were 295 µg/m3 and 146 µg/m3, and number concentrations were 140 000/cm3 and 100 000/cm3, respectively. Biomarkers were analyzed in samples of blood and urine taken before and several times after exposure. Results after wood smoke exposure were adjusted for exposure to filtered air. Results: Markers of systemic inflammation and soluble adhesion molecules did not increase after wood smoke exposure. Effects on markers of coagulation were ambiguous, with minor decreases in fibrinogen and platelet counts and mixed results concerning the coagulation factors VII and VIII. Urinary F2-isoprostane, a consistent marker of in vivo lipid peroxidation, unexpectedly decreased after wood smoke exposure. Conclusions: The effects on biomarkers of inflammation, coagulation and lipid peroxidation do not indicate an increased risk of cardiovascular diseases in healthy adults by short-term exposure to wood smoke at these moderate doses, previously shown to cause airway effects. PMID:23808634

  16. A 72 channel 125 MSPS analog-to-digital converter module for drift chamber readout for the GlueX detector

    SciTech Connect

    G. visser, D. Abbot, F. Barbosa, C. Cuevas, H. Dong, E. Jastrzembski, B. Moffitt, B. Raydo

    2011-06-01

    We have developed a very high channel density custom A/D converter module for the drift chamber readout requirements of the Glue-X detector at Jefferson Lab Hall-D. This 6U VME64× module has 72 channels, including a low noise differential line receiver, cable frequency response equalization, and signal shaping with a 5th order filter. The signals are digitized at 125 MSPS with an assembly option for either 12 or 14 bit resolution (GlueX will use 12 bit). The digitized data is written continuously to a circular buffer, with an independent read port and pipelined readout path able to extract potentially overlapping signal records at a 200 kHz trigger rate without deadtime. The extracted data is zero suppressed and may be further processed on-board to reduce the data volume, and is then assembled into event records and event block records. Up to 1 MB of output data is buffered on-board and available for readout via 2eSST at up to 320 MByte/s.

  17. SU-E-T-625: Use and Choice of Ionization Chambers for the Commissioning of Flattened and Flattening-Filter-Free Photon Beams: Determination of Recombination Correction Factor (ks)

    SciTech Connect

    Stucchi, C; Mongioj, V; Carrara, M; Pignoli, E; Bonfantini, F; Bresolin, A

    2014-06-15

    Purpose: To evaluate the recombination effect for some ionization chambers to be used for linacs commissioning for Flattened Filter (FF) and Flattening Filter Free (FFF) photon beams. Methods: A Varian TrueBeam linac with five photon beams was used: 6, 10 and 15 MV FF and 6 and 10 MV FFF. Measurements were performed in a water tank and in a plastic water phantom with different chambers: a mini-ion chamber (IC CC01, IBA), a plane-parallel ion chamber (IC PPC05, IBA) and two Farmer chambers (NE2581 and FPC05-IBA). Measurement conditions were Source- Surface Distance of 100 cm, two field sizes (10x10 and 40x40 cm2) and five depths (1cm, maximum buildup, 5cm, 10cm and 20cm). The ion recombination factors (kS), obtained from the Jaffe's plots (voltage interval 50-400 V), were evaluated at the recommended operating voltage of +300V. Results: Dose Per Pulse (DPP) at dmax was 0.4 mGy/pulse for FF beams, 1.0 mGy/pulse and 1.9 mGy/pulse for 6MV and 10 MV FFF beams respectively. For all measurement conditions, kS ranged between 0.996 and 0.999 for IC PPC05, 0.997 and 1.008 for IC CC01. For the FPC05 IBA Farmer IC, kS varied from 1.001 to 1.011 for FF beams, from 1.004 to 1.015 for 6 MV FFF and from 1.009 to 1.025 for 10 MV FFF. Whereas, for NE2581 IC the values ranged from 1.002 to 1.009 for all energy beams and measurement conditions. Conclusion: kS depends on the chamber volume and the DPP, which in turn depends on energy beam but is independent of dose rate. Ion chambers with small active volume can be reliably used for dosimetry of FF and FFF beams even without kS correction. On the contrary, for absolute dosimetry of FFF beams by Farmer ICs it is necessary to evaluate and apply the kS correction. Partially supported by Lega Italiana Lotta contro i Tumori (LILT)

  18. Bakeout Chamber Within Vacuum Chamber

    NASA Technical Reports Server (NTRS)

    Taylor, Daniel M.; Soules, David M.; Barengoltz, Jack B.

    1995-01-01

    Vacuum-bakeout apparatus for decontaminating and measuring outgassing from pieces of equipment constructed by mounting bakeout chamber within conventional vacuum chamber. Upgrade cost effective: fabrication and installation of bakeout chamber simple, installation performed quickly and without major changes in older vacuum chamber, and provides quantitative data on outgassing from pieces of equipment placed in bakeout chamber.

  19. Some recent trends in the evolution of gaseous detectors

    NASA Astrophysics Data System (ADS)

    Charpak, G.

    1982-05-01

    The emission of VUV light by electrons drifting in intense electric fields, with or without ionizing collisions, plays an important role in a variety of new classes of gaseous detectors, which are briefly analysed. New types of X-ray detectors with high-energy resolution, 8% fwhm at 6 keV, 1 mm spatial resolution, have been built. Large-surface VUV imaging photon detectors have important applications in Cherenkov ring imaging. Multistep avalanche chambers, invented for high-rate applications, appear to be a useful ingredient for single-photon detection, and find surprizing applications in applied fields such as high-accuracy chromatography or thermal neutron localization.

  20. Measurement of the ionization produced by sub-keV silicon nuclear recoils in a CCD dark matter detector

    DOE PAGES

    Chavarria, A. E.; Collar, J. I.; Peña, J. R.; ...

    2016-10-15

    We report a measurement of the ionization efficiency of silicon nuclei recoiling with sub-keV kinetic energy in the bulk silicon of a charge-coupled device (CCD). Nuclear recoils are produced by low-energy neutrons (<24 keV) from a 124Sb–9Be photoneutron source, and their ionization signal is measured down to 60 eV electron equivalent. This energy range, previously unexplored, is relevant for the detection of low-mass dark matter particles. The measured efficiency is found to deviate from the extrapolation to low energies of the Lindhard model. Furthermore, this measurement also demonstrates the sensitivity to nuclear recoils of CCDs employed by DAMIC, a darkmore » matter direct detection experiment located in the SNOLAB underground laboratory.« less

  1. Measurement of the ionization produced by sub-keV silicon nuclear recoils in a CCD dark matter detector

    SciTech Connect

    Chavarria, A. E.; Collar, J. I.; Peña, J. R.; Privitera, P.; Robinson, A. E.; Scholz, B.; Sengul, C.; Zhou, J.; Estrada, J.; Izraelevitch, F.; Tiffenberg, J.; de Mello Neto, J. R. T.; Machado, D. Torres

    2016-10-15

    We report a measurement of the ionization efficiency of silicon nuclei recoiling with sub-keV kinetic energy in the bulk silicon of a charge-coupled device (CCD). Nuclear recoils are produced by low-energy neutrons (<24 keV) from a 124Sb–9Be photoneutron source, and their ionization signal is measured down to 60 eV electron equivalent. This energy range, previously unexplored, is relevant for the detection of low-mass dark matter particles. The measured efficiency is found to deviate from the extrapolation to low energies of the Lindhard model. Furthermore, this measurement also demonstrates the sensitivity to nuclear recoils of CCDs employed by DAMIC, a dark matter direct detection experiment located in the SNOLAB underground laboratory.

  2. Detector to detector corrections: A comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams

    SciTech Connect

    Azangwe, Godfrey Grochowska, Paulina; Izewska, Joanna; Meghzifene, Ahmed; Georg, Dietmar; Hopfgartner, Johannes; Lechner, Wolfgang; Mizuno, Hideyuki; Fukumura, Akifumi; Yajima, Kaori; Gouldstone, Clare; Sharpe, Peter; Palmans, Hugo

    2014-07-15

    Purpose: The aim of the present study is to provide a comprehensive set of detector specific correction factors for beam output measurements for small beams, for a wide range of real time and passive detectors. The detector specific correction factors determined in this study may be potentially useful as a reference data set for small beam dosimetry measurements. Methods: Dose response of passive and real time detectors was investigated for small field sizes shaped with a micromultileaf collimator ranging from 0.6 × 0.6 cm{sup 2} to 4.2 × 4.2 cm{sup 2} and the measurements were extended to larger fields of up to 10 × 10 cm{sup 2}. Measurements were performed at 5 cm depth, in a 6 MV photon beam. Detectors used included alanine, thermoluminescent dosimeters (TLDs), stereotactic diode, electron diode, photon diode, radiophotoluminescent dosimeters (RPLDs), radioluminescence detector based on carbon-doped aluminium oxide (Al{sub 2}O{sub 3}:C), organic plastic scintillators, diamond detectors, liquid filled ion chamber, and a range of small volume air filled ionization chambers (volumes ranging from 0.002 cm{sup 3} to 0.3 cm{sup 3}). All detector measurements were corrected for volume averaging effect and compared with dose ratios determined from alanine to derive a detector correction factors that account for beam perturbation related to nonwater equivalence of the detector materials. Results: For the detectors used in this study, volume averaging corrections ranged from unity for the smallest detectors such as the diodes, 1.148 for the 0.14 cm{sup 3} air filled ionization chamber and were as high as 1.924 for the 0.3 cm{sup 3} ionization chamber. After applying volume averaging corrections, the detector readings were consistent among themselves and with alanine measurements for several small detectors but they differed for larger detectors, in particular for some small ionization chambers with volumes larger than 0.1 cm{sup 3}. Conclusions: The results demonstrate

  3. EVALUATION OF SMOKE AND GAS SENSOR RESPONSES FOR FIRES OF COMMON MINE COMBUSTIBLES

    PubMed Central

    Perera, Inoka Eranda; Litton, Charles D.

    2015-01-01

    Experiments were conducted to evaluate the response characteristics of commercially available gas, smoke, and flame sensors to fires of common combustible mine materials. The experiments were conducted in the large-scale Fire gallery located at the National Institute for Occupational Safety and Health (NIOSH) Lake Lynn Laboratory (LLL) in Fairchance, PA, using Ponderosa Pine, Red Oak, Douglas-fir, high and low volatile coals, PVC and SBR conveyor belt, No. 2 diesel fuel, and diesel exhaust. All the experiments (except those using No. 2 diesel fuel and the diesel exhaust tests) were conducted in a similar manner, with combustible materials heated rapidly by electrical strip heaters producing smoldering fires that quickly transitioned into flaming fires. The sensors included a diffusion-type carbon monoxide (CO) sensor, photoelectric- and ionization-type smoke sensors, a video smoke/flame detector, and an optical flame detector. Simultaneous measurements were obtained for average gas concentrations, smoke mass concentrations, and smoke optical densities in order to quantify the levels of combustion products at the alert and alarm times of the sensors. Because the required sensor alarm levels are 10 ppm and 0.044 m−1 optical density for CO and smoke sensors, respectively, the different sensor alarms are compared to the time at which the CO and smoke reached these alarm levels (1). In addition, the potential impact of using smoke sensors that have met the performance standards from accredited testing laboratories is also evaluated using the response of an Underwriters’ Laboratory (UL)-approved combination photoelectric/ionization smoke detector. The results are discussed relative to fire sensor needs that can have a positive impact on mine fire safety. PMID:26229418

  4. SU-C-304-01: Investigation of Various Detector Response Functions and Their Geometry Dependence in a Novel Method to Address Ion Chamber Volume Averaging Effect

    SciTech Connect

    Barraclough, B; Lebron, S; Li, J; Fan, Qiyong; Liu, C; Yan, G

    2015-06-15

    Purpose: A novel convolution-based approach has been proposed to address ion chamber (IC) volume averaging effect (VAE) for the commissioning of commercial treatment planning systems (TPS). We investigate the use of various convolution kernels and its impact on the accuracy of beam models. Methods: Our approach simulates the VAE by iteratively convolving the calculated beam profiles with a detector response function (DRF) while optimizing the beam model. At convergence, the convolved profiles match the measured profiles, indicating the calculated profiles match the “true” beam profiles. To validate the approach, beam profiles of an Elekta LINAC were repeatedly collected with ICs of various volumes (CC04, CC13 and SNC 125) to obtain clinically acceptable beam models. The TPS-calculated profiles were convolved externally with the DRF of respective IC. The beam model parameters were reoptimized using Nelder-Mead method by forcing the convolved profiles to match the measured profiles. We evaluated three types of DRFs (Gaussian, Lorentzian, and parabolic) and the impact of kernel dependence on field geometry (depth and field size). The profiles calculated with beam models were compared with SNC EDGE diode-measured profiles. Results: The method was successfully implemented with Pinnacle Scripting and Matlab. The reoptimization converged in ∼10 minutes. For all tested ICs and DRFs, penumbra widths of the TPS-calculated profiles and diode-measured profiles were within 1.0 mm. Gaussian function had the best performance with mean penumbra width difference within 0.5 mm. The use of geometry dependent DRFs showed marginal improvement, reducing the penumbra width differences to less than 0.3 mm. Significant increase in IMRT QA passing rates was achieved with the optimized beam model. Conclusion: The proposed approach significantly improved the accuracy of the TPS beam model. Gaussian functions as the convolution kernel performed consistently better than Lorentzian and

  5. Germanium detector vacuum encapsulation

    NASA Technical Reports Server (NTRS)

    Madden, N. W.; Malone, D. F.; Pehl, R. H.; Cork, C. P.; Luke, P. N.; Landis, D. A.; Pollard, M. J.

    1991-01-01

    This paper describes an encapsulation technology that should significantly improve the viability of germanium gamma-ray detectors for a number of important applications. A specialized vacuum chamber has been constructed in which the detector and the encapsulating module are processed in high vacuum. Very high vacuum conductance is achieved within the valveless encapsulating module. The detector module is then sealed without breaking the chamber vacuum. The details of the vacuum chamber, valveless module, processing, and sealing method are presented.

  6. A Laboratory Experiment To Measure Henry's Law Constants of Volatile Organic Compounds with a Bubble Column and a Gas Chromatography Flame Ionization Detector (GC-FID)

    ERIC Educational Resources Information Center

    Lee, Shan-Hu; Mukherjee, Souptik; Brewer, Brittany; Ryan, Raphael; Yu, Huan; Gangoda, Mahinda

    2013-01-01

    An undergraduate laboratory experiment is described to measure Henry's law constants of organic compounds using a bubble column and gas chromatography flame ionization detector (GC-FID). This experiment is designed for upper-division undergraduate laboratory courses and can be implemented in conjunction with physical chemistry, analytical…

  7. Development of a Liquefied Noble Gas Time Projection Chamber

    NASA Astrophysics Data System (ADS)

    Lesser, Ezra; White, Aaron; Aidala, Christine

    2015-10-01

    Liquefied noble gas detectors have been used for various applications in recent years for detecting neutrinos, neutrons, photons, and potentially dark matter. The University of Michigan is developing a detector with liquid argon to produce scintillation light and ionization electrons. Our data collection method will allow high-resolution energy measurement and spatial reconstruction of detected particles by using multi-pixel silicon photomultipliers (SiPM) and a cylindrical time projection chamber (TPC) with a multi-wire endplate. We have already designed a liquid argon condenser and purification unit surrounded by an insulating vacuum, constructed circuitry for temperature and pressure sensors, and created software to obtain high-accuracy sensor readouts. The status of detector development will be presented. Funded through the Michigan Memorial Phoenix Project.

  8. Focal-surface detector for heavy ions

    DOEpatents

    Erskine, John R.; Braid, Thomas H.; Stoltzfus, Joseph C.

    1979-01-01

    A detector of the properties of individual charged particles in a beam includes a gridded ionization chamber, a cathode, a plurality of resistive-wire proportional counters, a plurality of anode sections, and means for controlling the composition and pressure of gas in the chamber. Signals generated in response to the passage of charged particles can be processed to identify the energy of the particles, their loss of energy per unit distance in an absorber, and their angle of incidence. In conjunction with a magnetic spectrograph, the signals can be used to identify particles and their state of charge. The detector is especially useful for analyzing beams of heavy ions, defined as ions of atomic mass greater than 10 atomic mass units.

  9. Development of a high resolution liquid xenon imaging chamber for gamma-ray astronomy

    NASA Technical Reports Server (NTRS)

    Aprile, Elena

    1991-01-01

    The objective was to develop the technology of liquid xenon (LXe) detectors for spectroscopy and imaging of gamma rays from astrophysical sources emitting in the low to medium energy regime. In particular, the technical challenges and the physical processes relevant to the realization of the LXe detector operated as a Time Projection Chamber (TPC) were addressed and studied. Experimental results were obtained on the following topics: (1) long distance drift of free electrons in LXe (purity); (2) scintillation light yield for electrons and alphas in LXe (triggering); and (3) ionization yield for electrons and gamma rays in LXe (energy resolution). The major results from the investigations are summarized.

  10. Liquid-Argon Time Projection Chambers in the U.S

    SciTech Connect

    Soderberg, M.

    2009-10-01

    Liquid Argon Time Projection Chamber (LAr TPC) detectors are ideally suited for studying neutrino interactions and probing the parameters that characterize neutrino oscillations. The ability to drift ionization particles over long distances in purified argon and to trigger on abundant scintillation light allows for excellent particle identification and triggering capability. Recent U.S. based work in the development of LAr TPC technology for massive kiloton size detectors will be discussed in this talk, including details of the ArgoNeuT (Argon Neutrino Test) test-beam project, which is a 175 liter LAr TPC exposed to Fermilab's NuMI neutrino beamline.

  11. Exposure chamber

    DOEpatents

    Moss, Owen R.; Briant, James K.

    1983-01-01

    An exposure chamber includes an imperforate casing having a fluid inlet at the top and an outlet at the bottom. A single vertical series of imperforate trays is provided. Each tray is spaced on all sides from the chamber walls. Baffles adjacent some of the trays restrict and direct the flow to give partial flow back and forth across the chambers and downward flow past the lowermost pan adjacent a central plane of the chamber.

  12. Liquid Hydrogen: Target, Detector

    SciTech Connect

    Mulholland, G.T.; Harigel, G.G.

    2004-06-23

    In 1952 D. Glaser demonstrated that a radioactive source's radiation could boil 135 deg. C superheated-diethyl ether in a 3-mm O glass vessel and recorded bubble track growth on high-speed film in a 2-cm3 chamber. This Bubble Chamber (BC) promised improved particle track time and spatial resolution and cycling rate. Hildebrand and Nagle, U of Chicago, reported Liquid Hydrogen minimum ionizing particle boiling in August 1953. John Wood created the 3.7-cm O Liquid Hydrogen BC at LBL in January 1954. By 1959 the Lawrence Berkley Laboratory (LBL) Alvarez group's '72-inch' BC had tracks in liquid hydrogen. Within 10 years bubble chamber volumes increased by a factor of a million and spread to every laboratory with a substantial high-energy physics program. The BC, particle accelerators and special separated particle beams created a new era of High Energy Physics (HEP) experimentation. The BC became the largest most complex cryogenic installation at the world's HEP laboratories for decades. The invention and worldwide development, deployment and characteristics of these cryogenic dynamic target/detectors and related hydrogen targets are described.

  13. Development and Validation of a Method for Alcohol Analysis in Brain Tissue by Headspace Gas Chromatography with Flame Ionization Detector.

    PubMed

    Chun, Hao-Jung; Poklis, Justin L; Poklis, Alphonse; Wolf, Carl E

    2016-10-01

    Ethanol is the most widely used and abused drug. While blood is the preferred specimen for analysis, tissue specimens such as brain serve as alternative specimens for alcohol analysis in post-mortem cases where blood is unavailable or contaminated. A method was developed using headspace gas chromatography with flame ionization detection (HS-GC-FID) for the detection and quantification of ethanol, acetone, isopropanol, methanol and n-propanol in brain tissue specimens. Unfixed volatile-free brain tissue specimens were obtained from the Department of Pathology at Virginia Commonwealth University. Calibrators and controls were prepared from 4-fold diluted homogenates of these brain tissue specimens, and were analyzed using t-butanol as the internal standard. The chromatographic separation was performed with a Restek BAC2 column. A linear calibration was generated for all analytes (mean r(2) > 0.9992) with the limits of detection and quantification of 100-110 mg/kg. Matrix effect from the brain tissue was determined by comparing the slopes of matrix prepared calibration curves with those of aqueous calibration curves; no significant differences were observed for ethanol, acetone, isopropanol, methanol and n-propanol. The bias and the CVs for all volatile controls were ≤10%. The method was also evaluated for carryover, selectivity, interferences, bench-top stability and freeze-thaw stability. The HS-GC-FID method was determined to be reliable and robust for the analysis of ethanol, acetone, isopropanol, methanol and n-propanol concentrations in brain tissue, effectively expanding the specimen options for post-mortem alcohol analysis.

  14. Pulsed helium ionization detection system

    DOEpatents

    Ramsey, R.S.; Todd, R.A.

    1985-04-09

    A helium ionization detection system is provided which produces stable operation of a conventional helium ionization detector while providing improved sensitivity and linearity. Stability is improved by applying pulsed dc supply voltage across the ionization detector, thereby modifying the sampling of the detectors output current. A unique pulse generator is used to supply pulsed dc to the detector which has variable width and interval adjust features that allows up to 500 V to be applied in pulse widths ranging from about 150 nsec to about dc conditions.

  15. Pulsed helium ionization detection system

    DOEpatents

    Ramsey, Roswitha S.; Todd, Richard A.

    1987-01-01

    A helium ionization detection system is provided which produces stable operation of a conventional helium ionization detector while providing improved sensitivity and linearity. Stability is improved by applying pulsed dc supply voltage across the ionization detector, thereby modifying the sampling of the detectors output current. A unique pulse generator is used to supply pulsed dc to the detector which has variable width and interval adjust features that allows up to 500 V to be applied in pulse widths ranging from about 150 nsec to about dc conditions.

  16. Test chamber for alpha spectrometry

    DOEpatents

    Larsen, Robert P.

    1977-01-01

    Alpha emitters for low-level radiochemical analysis by measurement of alpha spectra are positioned precisely with respect to the location of a surface-barrier detector by means of a chamber having a removable threaded planchet holder. A pedestal on the planchet holder holds a specimen in fixed engagement close to the detector. Insertion of the planchet holder establishes an O-ring seal that permits the chamber to be pumped to a desired vacuum. The detector is protected against accidental contact and resulting damage.

  17. Separation and determination of benzene, toluene, ethylbenzene and o-xylene compounds in water using directly suspended droplet microextraction coupled with gas chromatography-flame ionization detector.

    PubMed

    Sarafraz-Yazdi, A; Amiri, A H; Es'haghi, Z

    2009-05-15

    The directly suspended droplet microextraction (DSDME) technique coupled with the capillary gas chromatography-flame ionization detector (GC-FID) was used to determine BTEX compounds in aqueous samples. The effective parameters such as organic solvent, extraction time, microdroplet volume, salt effect and stirring speed were optimized. The performance of the proposed technique was evaluated for the determination of BTEX compounds in natural water samples. Under the optimal conditions the enrichment factors ranged from 142.68 to 312.13, linear range; 0.01-20 microg mL(-1), limits of detection; 0.8-7 ng mL(-1) for most analytes. Relative standard deviations for 0.2 microg mL(-1) of BTEX in water were in the range 1.81-2.47% (n=5). The relative recoveries of BTEX from surface water at spiking level of 0.2 microg mL(-1) were in the range of 89.87-98.62%.

  18. A method for simultaneous analysis of phytosterols and phytosterol esters in tobacco leaves using non aqueous reversed phase chromatography and atmospheric pressure chemical ionization mass spectrometry detector.

    PubMed

    Ishida, Naoyuki

    2014-05-02

    While numerous analytical methods for phytosterols have been reported, the similar polarity and large molecules of phytosterol esters have made the methods lengthy and complicated. For this reason, an analytical method that could completely separate phytosterol esters including the higher fatty acids such as palmitic acid, stearic acid, oleic acid, linoleic acid and linolenic acid in addition to phytosterols without preliminary separation was developed. The separation was accomplished by non-aqueous reversed phase chromatography technique using only acetone and acetonitrile. An atmospheric pressure chemical ionization/mass spectrometry detector configured at selected ion monitoring mode was hyphenated with the separation system to detect phytosterols and phytosterol esters. Twenty-four types of these were consequently separated and then identified with their authentic components. The calibration curve was drawn in the range of about 5 to 25,000 ng/mL with a regression coefficient over 0.999. The limit of detection and limit of quantification, respectively, ranged from 0.9 to 3.0 ng/mL and from 3.0 to 11.0 ng/mL. Recovery rates ranged from 80 to 120%. The quantification results were subjected to statistical analysis and hierarchical clustering analysis, and were used to determine the differences in the amounts of phytosterols and phytosterol esters across tobacco leaves. The newly developed method succeeded in clarifying the whole composition of phytosterols and phytosterol esters in tobacco leaves and in explaining compositional differences across the variety of tobacco leaves.

  19. Lipids and Fatty Acids in Algae: Extraction, Fractionation into Lipid Classes, and Analysis by Gas Chromatography Coupled with Flame Ionization Detector (GC-FID).

    PubMed

    Guihéneuf, Freddy; Schmid, Matthias; Stengel, Dagmar B

    2015-01-01

    Despite the number of biochemical studies exploring algal lipids and fatty acid biosynthesis pathways and profiles, analytical methods used by phycologists for this purpose are often diverse and incompletely described. Potential confusion and potential variability of the results between studies can therefore occur due to change of protocols for lipid extraction and fractionation, as well as fatty acid methyl esters (FAME) preparation before gas chromatography (GC) analyses. Here, we describe a step-by-step procedure for the profiling of neutral and polar lipids using techniques such as solid-liquid extraction (SLE), thin-layer chromatography (TLC), and gas chromatography coupled with flame ionization detector (GC-FID). As an example, in this protocol chapter, analyses of neutral and polar lipids from the marine microalga Pavlova lutheri (an EPA/DHA-rich haptophyte) will be outlined to describe the distribution of fatty acid residues within its major lipid classes. This method has been proven to be a reliable technique to assess changes in lipid and fatty acid profiles in several other microalgal species and seaweeds.

  20. Characterization of a new commercial single crystal diamond detector for photon- and proton-beam dosimetry

    PubMed Central

    Akino, Yuichi; Gautam, Archana; Coutinho, Len; Würfel, Jan; Das, Indra J.

    2015-01-01

    A synthetic single crystal diamond detector (SCDD) is commercially available and is characterized for radiation dosimetry in various radiation beams in this study. The characteristics of the commercial SCDD model 60019 (PTW) with 6- and 15-MV photon beams, and 208-MeV proton beams, were investigated and compared with the pre-characterized detectors: Semiflex (model 31010) and PinPoint (model 31006) ionization chambers (PTW), the EDGE diode detector (Sun Nuclear Corp) and the SFD Stereotactic Dosimetry Diode Detector (IBA). To evaluate the effects of the pre-irradiation, the diamond detector, which had not been irradiated on the day, was set up in the water tank, and the response to 100 MU was measured every 20 s. The depth–dose and profiles data were collected for various field sizes and depths. For all radiation types and field sizes, the depth–dose data of the diamond chamber showed identical curves to those of the ionization chambers. The profile of the diamond detector was very similar to those of the EDGE and SFD detectors, although the Semiflex and PinPoint chambers showed volume-averaging effects in the penumbrae region. The temperature dependency was within 0.7% in the range of 4–41°C. A dose of 900 cGy and 1200 cGy was needed to stabilize the chamber to the level within 0.5% and 0.2%, respectively. The PTW type 60019 SCDD detector showed suitable characteristics for radiation dosimetry, for relative dose, depth–dose and profile measurements for a wide range of field sizes. However, at least 1000 cGy of pre-irradiation will be needed for accurate measurements. PMID:26268483

  1. Dosimetric properties of high energy current (HEC) detector in keV x-ray beams

    NASA Astrophysics Data System (ADS)

    Zygmanski, Piotr; Shrestha, Suman; Elshahat, Bassem; Karellas, Andrew; Sajo, Erno

    2015-04-01

    We introduce a new x-ray radiation detector. The detector employs high-energy current (HEC) formed by secondary electrons consisting predominantly of photoelectrons and Auger electrons, to directly convert x-ray energy to detector signal without externally applied power and without amplification. The HEC detector is a multilayer structure composed of thin conducting layers separated by dielectric layers with an overall thickness of less than a millimeter. It can be cut to any size and shape, formed into curvilinear surfaces, and thus can be designed for a variety of QA applications. We present basic dosimetric properties of the detector as function of x-ray energy, depth in the medium, area and aspect ratio of the detector, as well as other parameters. The prototype detectors show similar dosimetric properties to those of a thimble ionization chamber, which operates at high voltage. The initial results obtained for kilovoltage x-rays merit further research and development towards specific medical applications.

  2. Imaging nuclear decays with Optical Time Projection Chamber

    NASA Astrophysics Data System (ADS)

    Miernik, K.; Dominik, W.; Janas, Z.; Pfützner, M.; Bingham, C.; Czyrkowski, H.; Ćwiok, M.; Darby, I.; Dȧbrowski, R.; Fomitchev, A.; Gintei, T.; Golovkov, M.; Grzywacz, R.; Karny, M.; Korgul, A.; Kuśmierz, W.; Liddick, S.; Rajabali, M.; Rodin, A.; Rykaczewski, K.; Stepantsov, S.; Slepniev, R.; Stolz, A.; Ter-Akopian, G. M.; Wolski, R.

    2007-11-01

    A novel type of gaseous ionization detector—Optical Time Projection Chamber (OTPC)—developed to study rare nuclear decays is presented. The OTPC records tracks of charged particles ionizing a counting gas by optical imaging of the light generated by electrons multiplied in the amplification structures. By combining an electron drift-time profile measured by a photomultiplier and a CCD camera image we reconstruct three-dimensional trajectories of particles, energies and charges. The capabilities of the OTPC detector to study various decay modes are demonstrated by observation of beta-delayed proton emission from 13O, two-alpha break-up of 8Be, triple-alpha decay of 12C excited states and two-proton radioactivity of 45Fe.

  3. A Prototype Ionization Profile Monitor for RHIC.

    NASA Astrophysics Data System (ADS)

    Connolly, R.; Cameron, P.; Ryan, W.; Shea, T.; Sikora, R.; Tsoupas, N.

    1997-05-01

    Transverse beam profiles in the Relativistic Heavy-Ion Collider (RHIC) will be measured with ionization profile monitors (IPMs). Each IPM will measure the integrated distribution of electrons in one plane resulting from residual gas ionization during bunch passage. The high space-charge electric field of the beam makes it necessary to image with electrons which are guided by a magnetic field. A prototype detector was tested in the injection line during the RHIC Sextant Test. It consists of a collector circuit board mounted on one side of the beam and a parallel electrode on the other to provide an electric sweep field. The collector board has 48 electrodes oriented parallel to the beam with a chevron microchannel plate amplifier mounted in front of the collection traces. The detector vacuum chamber is placed in the gap of a magnet. At each bunch passage the charge pulses are integrated, amplified, and digitized for display as a profile histogram. This paper describes the prototype detector and gives results from the beam tests.

  4. Monitoring Liquid Argon Time Projection Chambers With A Raspberry Pi Camera

    NASA Astrophysics Data System (ADS)

    Patteson, Crystal

    2016-03-01

    The MicroBooNE detector is the first of three liquid argon (LAr) time projection chambers (TPCs) that are central to the short-baseline neutrino program at Fermilab. These chambers consist of thousands of stainless steel or beryllium-copper sense wires that detect ionization electrons produced when neutrinos interact with liquid argon nuclei inside the detector. The wires are several hundred microns in diameter to several meters in length. The construction of such LAr TPCs often takes place in an assembly hall, which is different from the detector hall where the experiment will operate, as was the case with MicroBooNE. Since in situ access to the chamber and its wires in the beamline enclosure can be limited, we investigate the possibility of using a Raspberry Pi single-board computer connected to a low-cost camera installed inside the cryostat as a cost-efficient way to verify the integrity of the wires after transport. We also highlight other benefits of this monitoring device implemented in MicroBooNE, including detector hall surveillance and verification of the status of LED indicators on detector electronics. The author would like to thank Dr. Matthew Toups for his encouragement and guidance on this research project.

  5. Analytical validation applied to simultaneous determination of solvents dichloromethane (DCM), methyl isobutyl ketone (MIBK), tetrahydrofuran (THF) and toluene (TOL) in urine by headspace extraction and injection on chromatographic system with a flame ionization detector

    NASA Astrophysics Data System (ADS)

    Muna, E. D. M.; Pereira, R. P.

    2016-07-01

    The determination of the volatile organic solvents dichloromethane (DCM), methyl isobutyl ketone (MIBK), tetrahydrofuran (THF) and toluene (TOL) is applied on toxicological monitoring of employees in various industrial activities. The gas chromatography technique with flame ionization detector and headspace injection system has been applied. The analytical procedure developed allows the simultaneous determination of the above-mentioned solvents and the accuracy of the method was tested following the INMETRO guidelines through the DOQ-CGRE 008 Rev.04-July/2011.

  6. DETECTORS AND EXPERIMENTAL METHODS: Study of a multi-wire proportional chamber with a cathode strip and delay-line readout

    NASA Astrophysics Data System (ADS)

    Han, Li-Ying; Li, Qi-Te; Faisal, Q.; Ge, Yu-Cheng; Liu, Hong-Tao; Ye, Yan-Lin

    2009-05-01

    The design principle for a multi-wire proportional chamber with a cathode strip and delay-line readout is described. A prototype chamber of a size of 10 cm ×10 cm was made together with the readout electronics circuit. A very clean signal with very low background noise was obtained by applying a transformer between the delay-line and the pre-amplifier in order to match the resistance. Along the anode wire direction a position resolution of less than 0.5 mm was achieved with a 55Fe-5.9 keV X ray source. The simple structure, large effective area and high position resolution allow the application of a gas chamber of this kind to many purposes.

  7. Exposure chamber

    DOEpatents

    Moss, Owen R.

    1980-01-01

    A chamber for exposing animals, plants, or materials to air containing gases or aerosols is so constructed that catch pans for animal excrement, for example, serve to aid the uniform distribution of air throughout the chamber instead of constituting obstacles as has been the case in prior animal exposure chambers. The chamber comprises the usual imperforate top, bottom and side walls. Within the chamber, cages and their associated pans are arranged in two columns. The pans are spaced horizontally from the walls of the chamber in all directions. Corresponding pans of the two columns are also spaced horizontally from each other. Preferably the pans of one column are also spaced vertically from corresponding pans of the other column. Air is introduced into the top of the chamber and withdrawn from the bottom. The general flow of air is therefore vertical. The effect of the horizontal pans is based on the fact that a gas flowing past the edge of a flat plate that is perpendicular to the flow forms a wave on the upstream side of the plate. Air flows downwardly between the chamber walls and the outer edges of the pan. It also flows downwardly between the inner edges of the pans of the two columns. It has been found that when the air carries aerosol particles, these particles are substantially uniformly distributed throughout the chamber.

  8. Radiation detector

    DOEpatents

    Fultz, B.T.

    1980-12-05

    Apparatus is provided for detecting radiation such as gamma rays and x-rays generated in backscatter Moessbauer effect spectroscopy and x-ray spectrometry, which has a large window for detecting radiation emanating over a wide solid angle from a specimen and which generates substantially the same output pulse height for monoenergetic radiation that passes through any portion of the detection chamber. The apparatus includes a substantially toroidal chamber with conductive walls forming a cathode, and a wire anode extending in a circle within the chamber with the anode lying closer to the inner side of the toroid which has the least diameter than to the outer side. The placement of the anode produces an electric field, in a region close to the anode, which has substantially the same gradient in all directions extending radially from the anode, so that the number of avalanche electrons generated by ionizing radiation is independent of the path of the radiation through the chamber.

  9. Radiation detector

    DOEpatents

    Fultz, Brent T.

    1983-01-01

    Apparatus is provided for detecting radiation such as gamma rays and X-rays generated in backscatter Mossbauer effect spectroscopy and X-ray spectrometry, which has a large "window" for detecting radiation emanating over a wide solid angle from a specimen and which generates substantially the same output pulse height for monoenergetic radiation that passes through any portion of the detection chamber. The apparatus includes a substantially toroidal chamber with conductive walls forming a cathode, and a wire anode extending in a circle within the chamber with the anode lying closer to the inner side of the toroid which has the least diameter than to the outer side. The placement of the anode produces an electric field, in a region close to the anode, which has substantially the same gradient in all directions extending radially from the anode, so that the number of avalanche electrons generated by ionizing radiation is independent of the path of the radiation through the chamber.

  10. A new method for direct total OH reactivity measurements using a fast Gas Chromatographic Photo-Ionization Detector (GC-PID)

    NASA Astrophysics Data System (ADS)

    Nölscher, A. C.; Sinha, V.; Bockisch, S.; Klüpfel, T.; Williams, J.

    2012-04-01

    *160*400 mm), light (8 kg) and provides a good time resolution (50-60 sec). The method is able to operate in ambient conditions, requires no external carrier gas, and an internal battery provides power for ca. 12 hours which makes the instrument suitable for outdoor field campaigns. The GC-PID detection limit for pyrrole is 3 ppb, which is adequate for this application and total OH reactivity can be measured under optimum conditions down to 4 sec-1. The new set-up has been tested in parallel to previously validated PTR-MS measurements for CRM in laboratory experiments, a plant chamber, and boreal forest field studies. Advantages and drawbacks of the new technique are discussed. Although the polymerization of pyrrole on the detector window remains a weakness in the present version of the system, in general the GC-PID produces acceptable results. [1] Calpini et al, 1999, Analusis, 27, 328 [2] Kovacs and Brune. 2001, Journal of Atmospheric Chemistry, 39, 105-122 [3] V. Sinha et al, 2008, Atmos. Chem. Phys., 8, 2213-2227 [4] http://www.environics-iut.de/

  11. Ionizing radiation from tobacco

    SciTech Connect

    Westin, J.B.

    1987-04-24

    Accidents at nuclear power facilities seem inevitably to bring in their wake a great deal of concern on the part of both the lay and medical communities. Relatively little attention, however, is given to what may be the largest single worldwide source of effectively carcinogenic ionizing radiation: tobacco. The risk of cancer deaths from the Chernobyl disaster are tobacco smoke is discussed.

  12. Determination of ionisation chamber collection efficiency in a swept electron beam by means of thermoluminescent detectors and the "two-voltage" method.

    PubMed

    Van Dam, J; Rijnders, A; Ang, K K; Mellaerts, M; Grobet, P

    1985-06-01

    Two methods for determining the collection efficiency of a 0.6 cm3 thimble ionisation chamber exposed to the swept electron beam of a linear accelerator Therac 20 Saturne (CGR MeV) have been compared. In one method the chamber signal has been compared to that of simultaneously exposed thermoluminescent LiF dosemeters (TLD), in the other the "two-voltage" method of Boag, adapted for swept beams, has been used. By variation of the electron energy between 20 and 13 MeV, of the focus-skin-distance (FSD) between 200 and 100 cm and of the monitor rate between 400 monitor units (m.u.) and 100 m.u. per minute, different values could be produced for the peak charge density M. The collection efficiency of the chamber, operating at a standard voltage of 250 V, decreases from 0.99 to 0.84 for a charge density increasing from 0.3 X 10(-4) C/m3 to 7.5 X 10(-4) C/m3, respectively. The maximum deviation observed between the TLD and the "two-voltage" method adopted for similar M is never more than 2% and mostly smaller than 1%. It can be concluded that, under the present experimental conditions, the calculated ionisation chamber collection efficiency is confirmed by the experimental method using TL dosimetry.

  13. Wire chamber

    DOEpatents

    Atac, Muzaffer

    1989-01-01

    A wire chamber or proportional counter device, such as Geiger-Mueller tube or drift chamber, improved with a gas mixture providing a stable drift velocity while eliminating wire aging caused by prior art gas mixtures. The new gas mixture is comprised of equal parts argon and ethane gas and having approximately 0.25% isopropyl alcohol vapor.

  14. Development, optimization, validation and application of faster gas chromatography - flame ionization detector method for the analysis of total petroleum hydrocarbons in contaminated soils.

    PubMed

    Zubair, Abdulrazaq; Pappoe, Michael; James, Lesley A; Hawboldt, Kelly

    2015-12-18

    This paper presents an important new approach to improving the timeliness of Total Petroleum Hydrocarbon (TPH) analysis in the soil by Gas Chromatography - Flame Ionization Detector (GC-FID) using the CCME Canada-Wide Standard reference method. The Canada-Wide Standard (CWS) method is used for the analysis of petroleum hydrocarbon compounds across Canada. However, inter-laboratory application of this method for the analysis of TPH in the soil has often shown considerable variability in the results. This could be due, in part, to the different gas chromatography (GC) conditions, other steps involved in the method, as well as the soil properties. In addition, there are differences in the interpretation of the GC results, which impacts the determination of the effectiveness of remediation at hydrocarbon-contaminated sites. In this work, multivariate experimental design approach was used to develop and validate the analytical method for a faster quantitative analysis of TPH in (contaminated) soil. A fractional factorial design (fFD) was used to screen six factors to identify the most significant factors impacting the analysis. These factors included: injection volume (μL), injection temperature (°C), oven program (°C/min), detector temperature (°C), carrier gas flow rate (mL/min) and solvent ratio (v/v hexane/dichloromethane). The most important factors (carrier gas flow rate and oven program) were then optimized using a central composite response surface design. Robustness testing and validation of model compares favourably with the experimental results with percentage difference of 2.78% for the analysis time. This research successfully reduced the method's standard analytical time from 20 to 8min with all the carbon fractions eluting. The method was successfully applied for fast TPH analysis of Bunker C oil contaminated soil. A reduced analytical time would offer many benefits including an improved laboratory reporting times, and overall improved clean up

  15. Application of portable gas chromatography-photo ionization detector combined with headspace sampling for field analysis of benzene, toluene, ethylbenzene, and xylene in soils.

    PubMed

    Zhou, You-Ya; Yu, Ji-Fang; Yan, Zeng-Guang; Zhang, Chao-Yan; Xie, Ya-Bo; Ma, Li-Qiang; Gu, Qing-Bao; Li, Fa-Sheng

    2013-04-01

    A method based on headspace (HS) sampling coupling with portable gas chromatography (GC) with photo ionization detector (PID) was developed for rapid determination of benzene, toluene, ethylbenzene, and xylenes (BTEX) in soils. Optimal conditions for HS gas sampling procedure were determined, and the influence of soil organic matter on the recovery of BTEX from soil was investigated using five representative Chinese soils. The results showed that the HS-portable-GC-PID method could be effectively operated at ambient temperature, and the addition of 15 ml of saturated NaCl solution in a 40-ml sampling vial and 60 s of shaking time for sample solution were optimum for the HS gas sampling procedure. The recoveries of each BTEX in soils ranged from 87.2 to 105.1 %, with relative standard deviations varying from 5.3 to 7.8 %. Good linearity was obtained for all BTEX compounds, and the detection limits were in the 0.1 to 0.8 μg kg(-1) range. Soil organic matter was identified as one of the principal elements that affect the HS gas sampling of BTEX in soils. The HS-portable-GC-PID method was successfully applied for field determination of benzene and toluene in soils of a former chemical plant in Jilin City, northeast China. Considering its satisfactory repeatability and reproducibility and particular suitability to be operated in ambient environment, HS sampling coupling with portable GC-PID is, therefore, recommended to be a suitable screening tool for rapid on-site determination of BTEX in soils.

  16. SYNCHROTRON RADIATION, FREE ELECTRON LASER, APPLICATION OF NUCLEAR TECHNOLOGY, ETC. Employing a Cerenkov detector for the thickness measurement of X-rays in a scattering background

    NASA Astrophysics Data System (ADS)

    Li, Shu-Wei; Kang, Ke-Jun; Wang, Yi; Li, Jin; Li, Yuan-Jing; Zhang, Qing-Jun

    2010-12-01

    The variation in environmental scattering background is a major source of systematic errors in X-ray inspection and measurement systems. As the energy of these photons consisting of environmental scattering background is much lower generally, the Cerenkov detectors having the detection threshold are likely insensitive to them and able to exclude their influence. A thickness measurement experiment is designed to verify the idea by employing a Cerenkov detector and an ionizing chamber for comparison. Furthermore, it is also found that the application of the Cerenkov detectors is helpful to exclude another systematic error from the variation of low energy components in the spectrum incident on the detector volume.

  17. Precise Measurement of Drift Velocities in Active-Target Detectors

    NASA Astrophysics Data System (ADS)

    Jensen, Louis

    2016-09-01

    Nuclear experiments with radioactive beams are needed to improve our understanding of nuclei structure far from stability. Radioactive beams typically have low beam rates, but active-target detectors can compensate for these low beam rates. In active-target detectors that are also Time-Projection Chambers (TPC), ionized electrons drift through an electric fieldto a detection device to imagethe trajectory of charged-particle ionization tracks within the chamber's gas volume. The measurement of the ionized electrons' drift velocity is crucial for the accurate imaging of these tracks. In order to measure this drift velocity, we will use a UV laser and photo-sensitive foil in a the ND-Cubedetector we are developing, periodically releasingelectrons from the foil at a known timesand a known distance from the electron detector, thereby precisely measuring the drift velocity in situ. We have surveyed several materials to find a material that will work well with typical solid-state UV lasers on the market. We plan to determine the best material and thickness of the foil to maximize the number of photoelectrons. The precision that will be afforded by this measurement of the drift velocity will allow us to eliminate a source of systematic uncertainty.

  18. Particle Morphology and Size Results from the Smoke Aerosol Measurement Experiment-2

    NASA Technical Reports Server (NTRS)

    Urban, David L.; Ruff, Gary A.; Greenberg, Paul S.; Fischer, David; Meyer, Marit; Mulholland, George; Yuan, Zeng-Guang; Bryg, Victoria; Cleary, Thomas; Yang, Jiann

    2012-01-01

    Results are presented from the Reflight of the Smoke Aerosol Measurement Experiment (SAME-2) which was conducted during Expedition 24 (July-September 2010). The reflight experiment built upon the results of the original flight during Expedition 15 by adding diagnostic measurements and expanding the test matrix. Five different materials representative of those found in spacecraft (Teflon, Kapton, cotton, silicone rubber and Pyrell) were heated to temperatures below the ignition point with conditions controlled to provide repeatable sample surface temperatures and air flow. The air flow past the sample during the heating period ranged from quiescent to 8 cm/s. The smoke was initially collected in an aging chamber to simulate the transport time from the smoke source to the detector. This effective transport time was varied by holding the smoke in the aging chamber for times ranging from 11 to 1800 s. Smoke particle samples were collected on Transmission Electron Microscope (TEM) grids for post-flight analysis. The TEM grids were analyzed to observe the particle morphology and size parameters. The diagnostics included a prototype two-moment smoke detector and three different measures of moments of the particle size distribution. These moment diagnostics were used to determine the particle number concentration (zeroth moment), the diameter concentration (first moment), and the mass concentration (third moment). These statistics were combined to determine the diameter of average mass and the count mean diameter and, by assuming a log-normal distribution, the geometric mean diameter and the geometric standard deviations can also be calculated. Overall the majority of the average smoke particle sizes were found to be in the 200 nm to 400 nm range with the quiescent cases producing some cases with substantially larger particles.

  19. Assessment of the setup dependence of detector response functions for mega-voltage linear accelerators

    SciTech Connect

    Fox, Christopher; Simon, Tom; Simon, Bill; Dempsey, James F.; Kahler, Darren; Palta, Jatinder R.; Liu Chihray; Yan Guanghua

    2010-02-15

    Purpose: Accurate modeling of beam profiles is important for precise treatment planning dosimetry. Calculated beam profiles need to precisely replicate profiles measured during machine commissioning. Finite detector size introduces perturbations into the measured profiles, which, in turn, impact the resulting modeled profiles. The authors investigate a method for extracting the unperturbed beam profiles from those measured during linear accelerator commissioning. Methods: In-plane and cross-plane data were collected for an Elekta Synergy linac at 6 MV using ionization chambers of volume 0.01, 0.04, 0.13, and 0.65 cm{sup 3} and a diode of surface area 0.64 mm{sup 2}. The detectors were orientated with the stem perpendicular to the beam and pointing away from the gantry. Profiles were measured for a 10x10 cm{sup 2} field at depths ranging from 0.8 to 25.0 cm and SSDs from 90 to 110 cm. Shaping parameters of a Gaussian response function were obtained relative to the Edge detector. The Gaussian function was deconvolved from the measured ionization chamber data. The Edge detector profile was taken as an approximation to the true profile, to which deconvolved data were compared. Data were also collected with CC13 and Edge detectors for additional fields and energies on an Elekta Synergy, Varian Trilogy, and Siemens Oncor linear accelerator and response functions obtained. Response functions were compared as a function of depth, SSD, and detector scan direction. Variations in the shaping parameter were introduced and the effect on the resulting deconvolution profiles assessed. Results: Up to 10% setup dependence in the Gaussian shaping parameter occurred, for each detector for a particular plane. This translated to less than a {+-}0.7 mm variation in the 80%-20% penumbral width. For large volume ionization chambers such as the FC65 Farmer type, where the cavity length to diameter ratio is far from 1, the scan direction produced up to a 40% difference in the shaping

  20. SU-E-T-82: A Study On Enhanced Dynamic Wedge (EDW) Dosimetry Using 2D Seven29 Ion Chamber Array Detector

    SciTech Connect

    Kumar, Syam; Aparna

    2015-06-15

    Purpose: To study the dosimetric properties of Enhanced Dynamic Wedge (EDW) using PTW Seven29 ion chamber array Methods: PTW Seven29 ion chamber array and Solid Water phantoms for different depths were used for the study. The study was carried out in Varian Clinac ix with photon energies, 6MV & 15MV. Primarily the solid water phantoms with the 2D array were scanned using a CT scanner (GE Optima 580) at different depths. These scanned images were used for EDW planning in an Eclipse treatment planning system (version 10). Planning was done for different wedge angles and for different depths for 6MV & 15MV. A dose of 100 CGy was delivered in each cases. For each delivery, calculated the Monitoring Unit (MU) required. Same set-up was created before delivering the plans in Varian Clinac-ix. For each clinically relevant depth and for different wedge angles, the same MU was delivered as calculated. Different wedged dose distributions where reconstructed from the measured 2D array data using the in-house developed excel program. Results: It is observed that the shoulder like region in the profile which reduces as depth increases. For the same depth and energy, the percentage difference between planned and measured dose is lesser than 3%. For smaller wedge angles, the percentage difference is found to be greater than 3% for the largest wedge angle. Standard deviation between measured doses at shoulder region for planned and measured profiles is 0.08 and 0.02 respectively. Standard deviations between planned and measured wedge factors for different depths (2.5cm, 5cm, 10cm, and 15cm) are (0.0021, 0.0007, 0.0050, 0.0001) for 6MV and (0.0024, 0.0191, 0.0013, 0.0005) for 15MV respectively. Conclusion: The 2D Seven29 ion chamber array is a good tool for the Enhanced Dynamic Wedge (EDW) dosimetry.

  1. Quantitation of Acrolein-Derived 3-Hydroxypropylmercapturic Acid in Human Urine by Liquid Chromatography-Atmospheric Pressure Chemical Ionization-Tandem Mass Spectrometry: Effects of Cigarette Smoking

    PubMed Central

    Carmella, Steven G.; Chen, Menglan; Zhang, Yan; Zhang, Siyi; Hatsukami, Dorothy K.; Hecht, Stephen S.

    2008-01-01

    Recently published data suggest that acrolein (1), a toxic but weakly carcinogenic constituent of cigarette smoke, may be involved as a causative factor for the mutations frequently observed in the p53 tumor suppressor gene in lung cancer in smokers. Biomarkers are needed to further assess the possible relationship between acrolein uptake and cancer. In this study, we analyzed 3-hydroxypropylmercapturic acid (3-HPMA, 2) in human urine. 3-HPMA is a major metabolite of acrolein in laboratory animals. The method employs [13C3]3-HPMA as internal standard, with analysis and quantitation by LC-APCI-MS/MS-SRM. Clean, readily quantifiable chromatograms were obtained. The method was accurate and precise and required only 0.1 mL of urine. Median levels of 3-HPMA were significantly higher (2900 pmol/mg creatinine, N = 35) in smokers than in non-smokers (683 pmol/mg creatinine, N = 21) (P = 0.0002). The effect of smoking was further assessed by determining levels of 3-HPMA before and after a 4 week smoking cessation period. There was a significant 78% decrease in median levels of urinary 3-HPMA after cessation (P < 0.0001). The relationship between levels of urinary 3-HPMA and those of acrolein-derived 1,N2-propanodeoxyguanosine (PdG) adducts in lung was investigated in 14 smokers. There was a significant inverse relationship between urinary 3-HPMA and α-hydroxy-PdG (3) but not γ-hydroxy-PdG (4) or total adduct levels. The results of this study clearly demonstrate that acrolein uptake in smokers is significantly higher than in non-smokers, and underline the need for further investigation of the possible relationship of acrolein uptake to lung cancer. PMID:17559234

  2. Ultrasonic liquid level detector

    DOEpatents

    Kotz, Dennis M.; Hinz, William R.

    2010-09-28

    An ultrasonic liquid level detector for use within a shielded container, the detector being tubular in shape with a chamber at its lower end into which liquid from in the container may enter and exit, the chamber having an ultrasonic transmitter and receiver in its top wall and a reflector plate or target as its bottom wall whereby when liquid fills the chamber a complete medium is then present through which an ultrasonic wave may be transmitted and reflected from the target thus signaling that the liquid is at chamber level.

  3. Automatic system for ionization chamber current measurements.

    PubMed

    Brancaccio, Franco; Dias, Mauro S; Koskinas, Marina F

    2004-12-01

    The present work describes an automatic system developed for current integration measurements at the Laboratório de Metrologia Nuclear of Instituto de Pesquisas Energéticas e Nucleares. This system includes software (graphic user interface and control) and a module connected to a microcomputer, by means of a commercial data acquisition card. Measurements were performed in order to check the performance and for validating the proposed design.

  4. Vertex detectors

    SciTech Connect

    Lueth, V.

    1992-07-01

    The purpose of a vertex detector is to measure position and angles of charged particle tracks to sufficient precision so as to be able to separate tracks originating from decay vertices from those produced at the interaction vertex. Such measurements are interesting because they permit the detection of weakly decaying particles with lifetimes down to 10{sup {minus}13} s, among them the {tau} lepton and charm and beauty hadrons. These two lectures are intended to introduce the reader to the different techniques for the detection of secondary vertices that have been developed over the past decades. The first lecture includes a brief introduction to the methods used to detect secondary vertices and to estimate particle lifetimes. It describes the traditional technologies, based on photographic recording in emulsions and on film of bubble chambers, and introduces fast electronic registration of signals derived from scintillating fibers, drift chambers and gaseous micro-strip chambers. The second lecture is devoted to solid state detectors. It begins with a brief introduction into semiconductor devices, and then describes the application of large arrays of strip and pixel diodes for charged particle tracking. These lectures can only serve as an introduction the topic of vertex detectors. Time and space do not allow for an in-depth coverage of many of the interesting aspects of vertex detector design and operation.

  5. The role of a microDiamond detector in the dosimetry of proton pencil beams.

    PubMed

    Gomà, Carles; Marinelli, Marco; Safai, Sairos; Verona-Rinati, Gianluca; Würfel, Jan

    2016-03-01

    In this work, the performance of a microDiamond detector in a scanned proton beam is studied and its potential role in the dosimetric characterization of proton pencil beams is assessed. The linearity of the detector response with the absorbed dose and the dependence on the dose-rate were tested. The depth-dose curve and the lateral dose profiles of a proton pencil beam were measured and compared to reference data. The feasibility of calibrating the beam monitor chamber with a microDiamond detector was also studied. It was found the detector reading is linear with the absorbed dose to water (down to few cGy) and the detector response is independent of both the dose-rate (up to few Gy/s) and the proton beam energy (within the whole clinically-relevant energy range). The detector showed a good performance in depth-dose curve and lateral dose profile measurements; and it might even be used to calibrate the beam monitor chambers-provided it is cross-calibrated against a reference ionization chamber. In conclusion, the microDiamond detector was proved capable of performing an accurate dosimetric characterization of proton pencil beams.

  6. Determination of 5alpha-androst-16-en-3alpha-ol in truffle fermentation broth by solid-phase extraction coupled with gas chromatography-flame ionization detector/electron impact mass spectrometry.

    PubMed

    Wang, Guan; Li, Yuan-Yuan; Li, Dong-Sheng; Tang, Ya-Jie

    2008-07-15

    A novel method using solid-phase extraction coupled with gas chromatography and flame ionization detector (FID)/electron impact mass spectrometry (EIMS) was developed for the determination of 5alpha-androst-16-en-3alpha-ol (androstenol), a steroidal compound belonging to the group of musk odorous 16-androstenes, in truffle fermentation broth. Comparison studies between FID and EIMS indicated two detectors gave similar quantitative results. The highest androstenol concentration of 123.5 ng/mL was detected in Tuber indicum fermentation broth, while no androstenol was found in Tuber aestivum fermentation broth. For the first time, this work confirmed the existence of androstenol in the truffle fermentation broth, which suggested truffle fermentation is a promising alternative for androstenol production on a large scale.

  7. An inexpensive dual-chamber particle monitor: laboratory characterization

    SciTech Connect

    Rufus Edwards; Kirk R. Smith; Brent Kirby; Tracy Allen; Charles D. Litton; Susanne Hering

    2006-06-15

    In developing countries, high levels of particle pollution from the use of coal and biomass fuels for household cooking and heating are a major cause of ill health and premature mortality. Existing monitoring equipment makes routine quantification of household particle pollution levels difficult. Recent advances have enabled the development of a small, portable, data-logging particle monitor modified from commercial smoke alarm technology that can meet the needs of surveys in the developing world at reasonable cost. Laboratory comparisons of a prototype particle monitor developed at the University of California at Berkeley (UCB) with gravimetric filters, a tapered element oscillating microbalance, and a TSI DustTrak to quantify the UCB particle monitor response as a function of both concentration and particle size and to examine sensor response in relation to changes in temperature, relative humidity, and elevation are presented. UCB particle monitors showed good linearity in response to different concentrations of laboratory-generated oleic acid aerosols with a coarse and fine size distributions. The photoelectric and ionization chamber showed a wide range of responses based on particle size and, thus, require calibration with the aerosol of interest. The ionization chamber was five times more sensitive to fine rather than coarse particles, whereas the photoelectric chamber was five times more sensitive to coarse than fine. The ratio of the response between the two sensors has the potential for mass calibration of individual data points based on estimated parameters of the size distribution. The results demonstrate the significant potential of this monitor, which will facilitate the evaluation of interventions (improved fuels, stoves, and ventilation) on indoor air pollution levels and research on the impacts of indoor particle levels on health in developing countries. 10 refs., 10 figs., 2 tabs.

  8. Multiwire proportional chamber development

    NASA Technical Reports Server (NTRS)

    Doolittle, R. F.; Pollvogt, U.; Eskovitz, A. J.

    1973-01-01

    The development of large area multiwire proportional chambers, to be used as high resolution spatial detectors in cosmic ray experiments is described. A readout system was developed which uses a directly coupled, lumped element delay-line whose characteristics are independent of the MWPC design. A complete analysis of the delay-line and the readout electronic system shows that a spatial resolution of about 0.1 mm can be reached with the MWPC operating in the strictly proportional region. This was confirmed by measurements with a small MWPC and Fe-55 X-rays. A simplified analysis was carried out to estimate the theoretical limit of spatial resolution due to delta-rays, spread of the discharge along the anode wire, and inclined trajectories. To calculate the gas gain of MWPC's of different geometrical configurations a method was developed which is based on the knowledge of the first Townsend coefficient of the chamber gas.

  9. The inverse problem of a Gaussian convolution and its application to the finite size of the measurement chambers/detectors in photon and proton dosimetry.

    PubMed

    Ulmer, W; Kaissl, W

    2003-03-21

    A Gaussian convolution kernel K is deduced as a Green's function of a Lie operator series. The deconvolution of a Gaussian kernel is developed by the inverse Green's function K(-1). A practical application is the deconvolution of measured profiles Dm(x) of photons and protons with finite detector size to determine the profiles Dp(x) of point-detectors or Monte Carlo Bragg curves of protons. The presented algorithms work if Dm(x) is either an analytical function or only given in a numerical form. Some approximation methods of the deconvolution are compared (differential operator expansion to analytical adaptations of 2 x 2 cm2 and 4 x 4 cm2 profiles, Hermite expansions to measured 6 x 6 cm2 and 20 x 20 cm2 profiles and Bragg curves of 80/180 MeV protons, FFT to an analytical 4 x 4 cm2 profile). The inverse problem may imply ill-posed problems, and, in particular, the use of FFT may be susceptible to them.

  10. Gaseous Detectors

    NASA Astrophysics Data System (ADS)

    Titov, Maxim

    Since long time, the compelling scientific goals of future high-energy physics experiments were a driving factor in the development of advanced detector technologies. A true innovation in detector instrumentation concepts came in 1968, with the development of a fully parallel readout for a large array of sensing elements - the Multi-Wire Proportional Chamber (MWPC), which earned Georges Charpak a Nobel prize in physics in 1992. Since that time radiation detection and imaging with fast gaseous detectors, capable of economically covering large detection volumes with low mass budget, have been playing an important role in many fields of physics. Advances in photolithography and microprocessing techniques in the chip industry during the past decade triggered a major transition in the field of gas detectors from wire structures to Micro-Pattern Gas Detector (MPGD) concepts, revolutionizing cell-size limitations for many gas detector applications. The high radiation resistance and excellent spatial and time resolution make them an invaluable tool to confront future detector challenges at the next generation of colliders. The design of the new micro-pattern devices appears suitable for industrial production. Novel structures where MPGDs are directly coupled to the CMOS pixel readout represent an exciting field allowing timing and charge measurements as well as precise spatial information in 3D. Originally developed for the high-energy physics, MPGD applications have expanded to nuclear physics, photon detection, astroparticle and neutrino physics, neutron detection, and medical imaging.

  11. Smoke and mirrors: a fiber optic smoke sensor

    NASA Astrophysics Data System (ADS)

    Whitesel, Henry K.; Overby, John K.; Ransford, Michael J.; Tatem, Patricia A.

    1994-11-01

    Smoke detectors in general, are usually threshold devices that frequently experience false alarms. Optical smoke detectors usually depend on the measurement of optical power absorption and scattering across an air gap and are usually threshold devices. Fiber optic sensor technology offers potential improvements for existing smoke detector technology. We have developed a new smoke sensor design based on wavelength selective absorption and scattering that generates a continuous measurement of smoke density. This technique provides first order compensation for water and dirt coatings on the optical surfaces and for optical power and ambient light changes. The sensor has a 2 inch sensing region and utilizes multimode technology with an 850 nanometer LED source. Experimental models of the fiber optic smoke sensors were tested successfully in our laboratory and on the ex-USS SHADWELL. Operational performance advantages of the fiber optic smoke sensor are expected in the areas of monitoring visibility, reducing false alarms, improving reliability, and continuous measurement of smoke density; this will improve fire detection capability and will assist in developing fire fighting strategy. Application of the sensors are planned for the shipboard environment to provide sensor input to new damage control management systems.

  12. An ionization profile monitor for the Tevatron

    SciTech Connect

    Jansson, A.; Bowden, M.; Bowie, K.; Bross, A.; Dysert, R.; Fitzpatrick, T.; Kwarciany, R.; Lundberg, C.; Nguyen, H.; Rivetta, C.; Slimmer, D.; Valerio, L.; Zagel, J.; /Fermilab

    2005-05-01

    Primarily to study emittance blowup during injection and ramping, an ionization profile monitor has been developed for the Tevatron. It is based on a prototype installed in the Main Injector, although with extensive modifications. In particular, the electromagnetic shielding has been improved, the signal path has been cleaned up, and provisions have been made for an internal electron source. Due to the good Tevatron vacuum, a local pressure bump is introduced to increase the primary signal, which is then amplified by a microchannel plate and detected on anode strips. For the DAQ, a custom ASIC developed for the CMS experiment is used. It is a combined charge integrator and digitizer, with a sensitivity of a few fC, and a time-resolution that allows single bunch measurement. Digitization is done in the tunnel to reduce noise. Preparations for detector installation were made during the long 2004 shutdown, with the installation of magnets, vacuum chambers, vacuum pumps and cabling. The actual detector will be installed during the fall 2005 shutdown. This paper describes the design of the detector and associated electronics, and presents various bench test results.

  13. Drift chamber tracking with neural networks

    SciTech Connect

    Lindsey, C.S.; Denby, B.; Haggerty, H.

    1992-10-01

    We discuss drift chamber tracking with a commercial log VLSI neural network chip. Voltages proportional to the drift times in a 4-layer drift chamber were presented to the Intel ETANN chip. The network was trained to provide the intercept and slope of straight tracks traversing the chamber. The outputs were recorded and later compared off line to conventional track fits. Two types of network architectures were studied. Applications of neural network tracking to high energy physics detector triggers is discussed.

  14. High-efficiency electron ionizer for a mass spectrometer array

    NASA Technical Reports Server (NTRS)

    Chutjian, Ara (Inventor); Darrach, Murray R. (Inventor); Orient, Otto J. (Inventor)

    2001-01-01

    The present invention provides an improved electron ionizer for use in a quadrupole mass spectrometer. The improved electron ionizer includes a repeller plate that ejects sample atoms or molecules, an ionizer chamber, a cathode that emits an electron beam into the ionizer chamber, an exit opening for excess electrons to escape, at least one shim plate to collimate said electron beam, extraction apertures, and a plurality of lens elements for focusing the extracted ions onto entrance apertures.

  15. Single wire drift chamber design

    SciTech Connect

    Krider, J.

    1987-03-30

    This report summarizes the design and prototype tests of single wire drift chambers to be used in Fermilab test beam lines. The goal is to build simple, reliable detectors which require a minimum of electronics. Spatial resolution should match the 300 ..mu..m rms resolution of the 1 mm proportional chambers that they will replace. The detectors will be used in beams with particle rates up to 20 KHz. Single track efficiency should be at least 99%. The first application will be in the MT beamline, which has been designed for calibration of CDF detectors. A set of four x-y modules will be used to track and measure the momentum of beam particles.

  16. Design and characterization of the beam monitor detectors of the Italian National Center of Oncological Hadron-therapy (CNAO)

    NASA Astrophysics Data System (ADS)

    Giordanengo, S.; Donetti, M.; Garella, M. A.; Marchetto, F.; Alampi, G.; Ansarinejad, A.; Monaco, V.; Mucchi, M.; Pecka, I. A.; Peroni, C.; Sacchi, R.; Scalise, M.; Tomba, C.; Cirio, R.

    2013-01-01

    A new hadron-therapy facility implementing an active beam scanning technique has been developed at the Italian National Center of Oncological Hadron-therapy (CNAO). This paper presents the design and the characterization of the beam monitor detectors developed for the on-line monitoring and control of the dose delivered during a treatment at CNAO. The detectors are based on five parallel-plate transmission ionization chambers with either a single large electrode or electrodes segmented in 128 strips (strip chambers) and 32×32 pixels (pixel chamber). The detectors are arranged in two independent boxes with an active area larger than 200×200 mm2 and a total water equivalent thickness along the beam path of about 0.9 mm. A custom front-end chip with 64 channels converts the integrated ionization channels without dead-time. The detectors were tested at the clinical proton beam facility of the Paul Scherrer Institut (PSI) which implements a spot scanning technique, each spot being characterized by a predefined number of protons delivered with a pencil beam in a specified point of the irradiation field. The short-term instability was measured by delivering several identical spots in a time interval of few tenths of seconds and is found to be lower than 0.3%. The non-uniformity, measured by delivering sequences of spots in different points of the detector surface, results to be lower than 1% in the single electrode chambers and lower than 1.5% in the strip and pixel chambers, reducing to less than 0.5% and 1% in the restricted 100×100 mm2 central area of the detector.

  17. Overview of the n3He Experiment and Target Chamber

    NASA Astrophysics Data System (ADS)

    McCrea, Mark; n3He Collaboration

    2017-01-01

    The n3He Experiment aims to measure the parity-violating asymmetry in the direction of proton emission relative to the initial neutron polarization direction in the reaction n-> +3 He -> T + p + 765 keV to a high precision. The size of the asymmetry is estimated to be in the range - 9 . 5 - 2 . 5 ×10-8 , and our goal statistical accuracy is 2 ×10-8 . The experiment ran at the Spallation Neutron Source with data taking completing at the end of 2015. The experiment used a Helium-3 ionization chamber as the combined target and detector. Data analysis is underway and is currently in an advanced stage

  18. The Japanese Radon and Thoron Reference Chambers

    SciTech Connect

    Tokonami, Shinji; Ishikawa, Tetsuo; Sorimachi, Atsuyuki; Takahashi, Hiroyuki; Miyahara, Nobuyuki

    2008-08-07

    Passive detectors used for large-scale and long-term surveys are generally calibrated in a well-controlled environment such as a radon chamber. It has been also pointed out that some of them are sensitive to thoron. Thus it is necessary to check the thoron contribution to the detector response with the proposed or similar test before practical use. The NIRS accommodates radon/aerosol and thoron chambers for quality assurance and quality control of radon measurements. Thus both chambers work so well that they can supply us with the calibration technique and consequently, a good level of knowledge of the radon and thoron issue.

  19. Calibration of the active radiation detector for Spacelab-One

    NASA Technical Reports Server (NTRS)

    1982-01-01

    The flight models of the active radiation detector (ARD) for the ENV-01 environmental monitor were calibrated using gamma radiation. Measured sensitivities of the ion chambers were 6.1 + or - 0.3 micron rad per count for ARD S/N1, and 10.4 + or - 0.5 micron rad per count for ARD S/N2. Both were linear over the measured range 0.10 to 500 m/rad hour. The particle counters (proportional counters) were set to respond to approximately 85% of minimum ionizing particles of unit charge passing through them. These counters were also calibrated in the gamma field.

  20. Development of 10B-Based 3He Replacement Neutron Detectors

    NASA Astrophysics Data System (ADS)

    King, Michael J.; Gozani, Tsahi; Hilliard, Donald B.

    2011-12-01

    Radiation portal monitors (RPM) are currently deployed at United States border crossings to passively inspect vehicles and persons for any emission of neutrons and/or gamma rays, which may indicate the presence of unshielded nuclear materials. The RPM module contains an organic scintillator with 3He proportional counters to detect gamma rays and thermalized neutrons, respectively. The supply of 3He is rapidly dwindling, requiring alternative detectors to provide the same function and performance. Our alternative approach is one consisting of a thinly-coated 10B flat-panel ionization chamber neutron detector that can be deployed as a direct drop-in replacement for current RPM 3He detectors. The uniqueness of our approach in providing a large-area detector is in the simplicity of construction, scalability of the unit cell detector, ease of adaptability to a variety of applications and low cost. Currently, Rapiscan Laboratories and Helicon Thin Film Systems have designed and developed an operational 100 cm2 multi-layer prototype 10BB-based ionization chamber.

  1. Comparison measurements of DQE for two flat panel detectors: fluoroscopic detector vs. cone beam CT detector

    NASA Astrophysics Data System (ADS)

    Betancourt Benítez, Ricardo; Ning, Ruola; Conover, David

    2006-03-01

    The physical performance of two flat panel detectors (FPD) has been evaluated using a standard x-ray beam quality set by IEC, namely RQA5. The FPDs evaluated in this study are based on an amorphous silicon photodiode array that is coupled to a thallium-doped Cesium Iodide scintillator and to a thin film transistor (TFT) array. One detector is the PaxScan 2520 that is designed for fluoro imaging, and has a small dynamic range and a large image lag. The other detector is the PaxScan 4030CB that is designed for cone beam CT, and has a large dynamic range (>16-bit), a reduced image lag and many imaging modes. Varian Medical Systems manufactured both detectors. The linearity of the FPDs was investigated by using an ionization chamber and aluminum filtration in order to obtain the beam quality. Since the FPDs are used in fluoroscopic mode, image lag of the FPD was measured in order to investigate its effect on this study, especially its effect on DQE. The spatial resolution of the FPDs was determined by obtaining the pre-sampling modulation transfer function for each detector. A sharp edge was used in accordance to IEC 62220-1. Next, the Normalized Noise Power Spectrum (NNPS) was calculated for various exposures levels at RQA5 radiation quality. Finally, the DQE of each FPD was obtained with a modified version of the international standard set by IEC 62220-1. The results show that the physical performance in DQE and MTF of the PaxScan 4030CB is superior to that of PaxScan2520.

  2. Development of bubble chambers with enhanced stability and sensitivity to low-energy nuclear recoils

    SciTech Connect

    Bolte, W.J.; Collar, Juan I.; Crisler, M.; Hall, J.; Holmgren, D.; Nakazawa, D.; Odom, B.; O'Sullivan, K.; Plunkett, R.; Ramberg, E.; Raskin, A.; Sonnenschein, A.; Vieira, J.D.; /Chicago U., EFI /KICP, Chicago /Fermilab

    2005-03-01

    The viability of using a Bubble Chamber for rare event searches and in particular for the detection of dark matter particle candidates is considered. Techniques leading to the deactivation of inhomogeneous nucleation centers and subsequent enhanced stability in such a detector are described. Results from prototype trials indicate that sensitivity to low-energy nuclear recoils like those expected from Weakly Interacting Massive Particles can be obtained in conditions of near total insensitivity to minimum ionizing backgrounds. An understanding of the response of superheated heavy refrigerants to these recoils is demonstrated within the context of existing theoretical models. We comment on the prospects for the detection of supersymmetric dark matter particles with a large CF{sub 3}I chamber.

  3. Comments on GUT monopole energy loss and ionization

    SciTech Connect

    Hagstrom, R.

    1982-01-01

    A few comments about the likely behavior of the electromagnetic energy loss and ionization rates of super-slowly moving magnetic monopoles are presented. The questions of energy loss rates and ionization rates for super-low monopoles passing through matter are considered, concentrating on aspects of these issues which affect practical detection techniques. It is worthwhile here to emphasize that there is a potentially great distinction between energy loss rates and ionization rates and that the magnitude of this distinction is really the great issue which must be settled in order to understand the significance of experimental results from present and proposed investigations of the slow monopole question. Energy loss here means the total dE/dX of the projectile due to interactions with the electrons of the slowing medium. To the extent that nuclear collisions can be neglected, this so-called electronic energy loss is the relevant quantity in questions about whether monopoles stop within the earth's crust, whether they are slowed by interstellar plasmas, or the signal in a truly calorimetric measurement (measuring temperature rises along the trajectory), etc. Most of our successful detection techniques depend upon the promotion of ground state electrons into states which lie above some energy gap in the material of the detector: electrons must be knocked completely free from the gas atoms in a proportional chamber gas, electrons must be promoted to a higher band in solid scintillator plastics. These processes are generically identified as ionization. (WHK)

  4. Photon counting detector for the personal radiography inspection system "SIBSCAN"

    NASA Astrophysics Data System (ADS)

    Babichev, E. A.; Baru, S. E.; Grigoriev, D. N.; Leonov, V. V.; Oleynikov, V. P.; Porosev, V. V.; Savinov, G. A.

    2017-02-01

    X-ray detectors operating in the energy integrating mode are successfully used in many different applications. Nevertheless the direct photon counting detectors, having the superior parameters in comparison with the integrating ones, are rarely used yet. One of the reasons for this is the low value of the electrical signal generated by a detected photon. Silicon photomultiplier (SiPM) based scintillation counters have a high detection efficiency, high electronic gain and compact dimensions. This makes them a very attractive candidate to replace routinely used detectors in many fields. More than 10 years ago the digital scanning radiography system based on multistrip ionization chamber (MIC) was suggested at Budker Institute of Nuclear Physics. The detector demonstrates excellent radiation resistance and parameter stability after 5 year operations and an imaging of up to 1000 persons per day. Currently, the installations operate at several Russian airports and at subway stations in some cities. At the present time we design a new detector operating in the photon counting mode, having superior parameters than the gas one, based on scintillator - SiPM assemblies. This detector has close to zero noise, higher quantum efficiency and a count rate capability of more than 5 MHz per channel (20% losses), which leads to better image quality and improved detection capability. The suggested detector technology could be expanded to medical applications.

  5. Experimental study of the rigidity and transparency to ionizing radiation of composite materials used in the enclosure under pressure of the Micromegas detector

    NASA Astrophysics Data System (ADS)

    Harbaoui, Imen; Besbes, Hatem; Chafra, Moez

    Innovation in the field of nuclear imaging is necessarily followed by a radical change in the detection principle. The gas detector Micromegas (Mesh Micro Structure Gaseous) could be an interesting option, thanks to the stability and robustness of such a detector. Thus, it was necessary to study the implementation of the detector enclosure in composite materials. The focus of the present study was the robustness and gamma rays transparency of a set of composites. The studied composites were reinforced with vegetable fibers (alfa), and synthetic fibers. The mechanical properties of all composites specimen were evaluated by three-point bending test, whereas, gamma ray transparency was evaluated by the exposition of composites specimen to a mono-energetic gamma ray beam emitted by a Technetium 99-m source. Findings revealed that the biocomposite materials using alfa fiber and Polymethyl Methacrylate matrix are very promising as long as they present good robustness and high gamma ray transparency in diagnostic range.

  6. The Mark II Vertex Drift Chamber

    SciTech Connect

    Alexander, J.P.; Baggs, R.; Fujino, D.; Hayes, K.; Hoard, C.; Hower, N.; Hutchinson, D.; Jaros, J.A.; Koetke, D.; Kowalski, L.A.

    1989-03-01

    We have completed constructing and begun operating the Mark II Drift Chamber Vertex Detector. The chamber, based on a modified jet cell design, achieves 30 {mu}m spatial resolution and <1000 {mu}m track-pair resolution in pressurized CO{sub 2} gas mixtures. Special emphasis has been placed on controlling systematic errors including the use of novel construction techniques which permit accurate wire placement. Chamber performance has been studied with cosmic ray tracks collected with the chamber located both inside and outside the Mark II. Results on spatial resolution, average pulse shape, and some properties of CO{sub 2} mixtures are presented. 10 refs., 12 figs., 1 tab.

  7. Design and First Results of the CoDeX Liquid-Xenon Compton-Imaging Detector

    NASA Astrophysics Data System (ADS)

    Tennyson, Brian; Cahn, Sidney; Bernard, Ethan; Boulton, Elizabeth; Destefano, Nicholas; Edwards, Blair; Hackenburg, Ariana; Horn, Markus; Larsen, Nicole; Nikkel, James; Wahl, Christopher; Gai, Moshe; McKinsey, Daniel

    2016-03-01

    CoDeX (Compton-imaging Detector in Xenon) is an R&D Compton gamma-ray imaging detector that uses 30 kg of xenon in a two-phase time projection chamber. Time projection relative to the initial scintillation signal provides the vertical interaction positions, and either PMT-sensed gas electroluminescence or a charge-sensitive amplifier quantifies the drifted ionization signal. Detector features to enable Compton imaging are a pair of instrumented wire grids added to sense the horizontal position of clouds of drifted electrons that traverse the detector. Each wire is individually amplified in the cold xenon environment. Design choices addressing the thermodynamic and xenon purity constraints of this system will be discussed. We will also discuss the mechanical designs, engineering challenges, and performance of this Compton-imaging detector.

  8. Smoking Cessation

    MedlinePlus

    ... National Center for Chronic Disease Prevention and Health Promotion, Office on Smoking and Health, 2014 [accessed 2017 ... National Center for Chronic Disease Prevention and Health Promotion, Office on Smoking and Health, 2010 [accessed 2017 ...

  9. Secondhand Smoke

    MedlinePlus

    ... National Center for Chronic Disease Prevention and Health Promotion, Office on Smoking and Health, 2014 [accessed 2017 ... National Center for Chronic Disease Prevention and Health Promotion, Office on Smoking and Health, 2010 [accessed 2017 ...

  10. Wood Smoke

    EPA Pesticide Factsheets

    Smoke is made up of a complex mixture of gases and fine, microscopic particles produced when wood and other organic matter burn. The biggest health threat from wood smoke comes from fine particles (also called particulate matter).

  11. Asia Smoke

    Atmospheric Science Data Center

    2013-04-16

    ... title:  Smoke from Asian Fires Traverses the Pacific     View Larger Image ... moved eastwards over the northern portion of the Pacific Ocean, the thickness of the smoke passing over an area south of the Aleutian ...

  12. Teen Smoking

    MedlinePlus

    ... or product placement in movies that create the perception that smoking is glamorous and more prevalent than ... org/healthy-lifestyle/tween-and-teen-health/in-depth/teen-smoking/art-20047069 . Mayo Clinic Footer Legal ...

  13. It might work in Oklahoma but will it work in Oakhampton? Context and implementation in the effectiveness literature on domestic smoke detectors

    PubMed Central

    Arai, L; Roen, K; Roberts, H; Popay, J

    2005-01-01

    Objective: To explore data on factors affecting implementation processes in papers contributing to a Cochrane systematic review (SR) of smoke alarm interventions, supplemented by further papers not included in the review. Design: Screening for data on implementation on the basis of: (1) primary studies included in a Cochrane SR, (2) further papers relating to these and similar studies, and (3) approaches to authors of these and other relevant studies and reports. Results: Relatively few data were found to help people seeking to implement smoke alarm promotion interventions. Conclusions: For practitioners and policymakers to be able to build on research evidence, researchers and journal editors need to ensure that sufficient data are published, or are otherwise available to interested parties to move from understanding the evidence to using it. PMID:15933405

  14. Infrared Optical Readout of a Gas-Based Recoil Tracking Detector

    NASA Astrophysics Data System (ADS)

    Miller, Katrina; Barbeau, Phillip; Rich, Grayson; Awe, Connor

    2016-03-01

    Gas-based recoil tracking detectors are used in a variety of nuclear and particle physics experiments to identify particles based on distinct interaction signatures. Past research shows that this technology, if further developed, may prove useful in the ongoing search for dark matter and coherent neutrino scattering observations. This research presents the original design and development of a tracking detector that uses gaseous argon as a scintillating material to measure infrared optical readout. The initial model of this detector, consisting of a wire chamber filled with P-10, has produced unambiguous ionization signals. Current studies are focused toward using pure gaseous argon to detect coincident scintillation signals, which will demonstrate the capability of the detector to image particle tracks using nonvisible radiation.

  15. Fire detector response in aircraft applications

    NASA Technical Reports Server (NTRS)

    Wiersma, S. J.; Mckee, R. G.

    1978-01-01

    Photoelectric, ionization, and gas sensors were used to detect the signatures from the radiant heat or flame of various aircraft materials. It was found that both ionization and photoelectric detectors are about equally capable of detecting products of pyrolysis and combustion of synthetic polymers, especially those containing fire-retardant additives. Ionization detectors alone appeared to be sensitive to combustion products of simple cellulosic materials. A gas sensor detector appeared to be insensitive to pyrolysis or combustion products of many of the materials.

  16. A two-dimensional position sensitive gas chamber with scanned charge transfer readout

    NASA Astrophysics Data System (ADS)

    Gómez, F.; Iglesias, A.; Lobato, R.; Mosquera, J.; Pardo, J.; Pena, J.; Pazos, A.; Pombar, M.; Rodríguez, A.

    2003-10-01

    We have constructed and tested a two-dimensional position sensitive parallel-plate gas ionization chamber with scanned charge transfer readout. The scan readout method described here is based on the development of a new position-dependent charge transfer technique. It has been implemented by using gate strips perpendicularly oriented to the collector strips. This solution reduces considerably the number of electronic readout channels needed to cover large detector areas. The use of a 25 μm thick kapton etched circuit allows high charge transfer efficiency with a low gating voltage, consequently needing a very simple commutating circuit. The present prototype covers 8×8 cm2 with a pixel size of 1.27×1.27 mm2. Depending on the intended use and beam characteristics a smaller effective pixel is feasible and larger active areas are possible. This detector can be used for X-ray or other continuous beam intensity profile monitoring.

  17. Automobile dirty smoke eliminator

    SciTech Connect

    Chen, H.

    1981-08-04

    An automobile dirty smoke eliminator is disclosed which mainly consists of two oval tin plates externally, and upper and lower exhaust pipes , left and right support plates, a blade, a discharge chamber, a discharge pipe, etc. internally. The principle of the present invention mainly consists in making use of the effect of mixing water and gas to entirely eliminate automobile dirty smoke. When the dirty smoke (exhaust gas) of automobile enters into the lower exhaust pipe of the present invention, the blade at the outlet of said lower exhaust pipe submerged in water is impacted by the compression force derived from the engine exhaust stroke so as to generate a mixture of said water and exhaust gas and to form a whirlpool having many buddles. The effect of walls of left and right support plates promote the toxin in the said dirty smoke (Exhaust gas) automatically to deposit in the said water. The surplus toxin is discharged through the upper exhaust pipe and then mixed with the water to form a colorless, odorless nontoxious fog-like vapor which exhausts through the discharge pipe in order to achieve the best result of eliminating the said dirty smoke (Exhaust gas).

  18. Quitting Smoking

    MedlinePlus

    Tobacco use is the most common preventable cause of death. About half of the people who don't quit smoking will die of smoking-related problems. Quitting smoking is important for your health. Soon after you quit, ... In the long term, giving up tobacco can help you live longer. Your risk of ...

  19. Atmospheric sampling glow discharge ionization source

    DOEpatents

    McLuckey, S.A.; Glish, G.L.

    1989-07-18

    An atmospheric sampling glow discharge ionization source that can be used in combination with an analytical instrument which operates at high vacuum, such as a mass spectrometer. The atmospheric sampling glow discharge ionization source comprises a chamber with at least one pair of electrodes disposed therein, an inlet for a gaseous sample to be analyzed and an outlet communicating with an analyzer which operates at subatmospheric pressure. The ionization chamber is maintained at a pressure below atmospheric pressure, and a voltage difference is applied across the electrodes to induce a glow discharge between the electrodes, so that molecules passing through the inlet are ionized by the glow discharge and directed into the analyzer. The ionization source accepts the sample under atmospheric pressure conditions and processes it directly into the high vacuum instrument, bridging the pressure gap and drawing off unwanted atmospheric gases. The invention also includes a method for analyzing a gaseous sample using the glow discharge ionization source described above. 3 figs.

  20. Performance studies of Micro Pixel Chamber for the ATLAS upgrade

    NASA Astrophysics Data System (ADS)

    Komai, H.; Ochi, A.; Homma, Y.; Edo, Y.; Yamaguchi, T.

    2013-03-01

    The Micro Pixel Chamber (μ-PIC) is being developed as a muon chamber of the ATLAS experiment in an HL-LHC environment. In the ATLAS muon system, a high flux of fast neutron background causes instability operation of the detectors. We performed neutron irradiation tests with μ-PIC to optimize the operation gas and detector structure. In addition, we studied neutron interactions with the detector, in order to understand the effect of fast neutrons.

  1. Diogene pictorial drift chamber

    SciTech Connect

    Gosset, J.

    1984-01-01

    A pictorial drift chamber, called DIOGENE, has been installed at Saturne in order to study central collisions of high energy heavy ions. It has been adapted from the JADE internal detector, with two major differences to be taken into account. First, the center-of-mass of these collisions is not identical to the laboratory reference frame. Second, the energy loss and the momentum ranges of the particles to be detected are different from the ones in JADE. It was also tried to keep the cost as small as possible, hence the choice of minimum size and minimum number of sensitive wires. Moreover the wire planes are shifted from the beam axis: this trick helps very much to quickly reject the bad tracks caused by the ambiguity of measuring drift distances (positive or negative) through times (always positive).

  2. Detecting dark matter with scintillating bubble chambers

    NASA Astrophysics Data System (ADS)

    Zhang, Jianjie; Dahl, C. Eric; Jin, Miaotianzi; Baxter, Daniel

    2016-03-01

    Threshold based direct WIMP dark matter detectors such as the superheated bubble chambers developed by the PICO experiment have demonstrated excellent electron-recoil and alpha discrimination, excellent scalability, ease of change of target fluid, and low cost. However, the nuclear-recoil like backgrounds have been a limiting factor in their dark matter sensitivity. We present a new type of detector, the scintillating bubble chamber, which reads out the scintillation pulse of the scattering events as well as the pressure, temperature, acoustic traces, and bubble images as a conventional bubble chamber does. The event energy provides additional handle to discriminate against the nuclear-recoil like backgrounds. Liquid xenon is chosen as the target fluid in our prototyping detector for its high scintillation yield and suitable vapor pressure which simplifies detector complexity. The detector can be used as an R&D tool to study the backgrounds present in the current PICO bubble chambers or as a prototype for standalone dark matter detectors in the future. Supported by DOE Grant DE-SC0012161.

  3. Concepts for dose determination in flat-detector CT

    NASA Astrophysics Data System (ADS)

    Kyriakou, Yiannis; Deak, Paul; Langner, Oliver; Kalender, Willi A.

    2008-07-01

    Flat-detector computed tomography (FD-CT) scanners provide large irradiation fields of typically 200 mm in the cranio-caudal direction. In consequence, dose assessment according to the current definition of the computed tomography dose index CTDIL=100 mm, where L is the integration length, would demand larger ionization chambers and phantoms which do not appear practical. We investigated the usefulness of the CTDI concept and practical dosimetry approaches for FD-CT by measurements and Monte Carlo (MC) simulations. An MC simulation tool (ImpactMC, VAMP GmbH, Erlangen, Germany) was used to assess the dose characteristics and was calibrated with measurements of air kerma. For validation purposes measurements were performed on an Axiom Artis C-arm system (Siemens Medical Solutions, Forchheim, Germany) equipped with a flat detector of 40 cm × 30 cm. The dose was assessed for 70 kV and 125 kV in cylindrical PMMA phantoms of 160 mm and 320 mm diameter with a varying phantom length from 150 to 900 mm. MC simulation results were compared to the values obtained with a calibrated ionization chambers of 100 mm and 250 mm length and to thermoluminesence (TLD) dose profiles. The MCs simulations were used to calculate the efficiency of the CTDIL determination with respect to the desired CTDI∞. Both the MC simulation results and the dose distributions obtained by MC simulation were in very good agreement with the CTDI measurements and with the reference TLD profiles, respectively, to within 5%. Standard CTDI phantoms which have a z-extent of 150 mm underestimate the dose at the center by up to 55%, whereas a z-extent of >=600 mm appears to be sufficient for FD-CT; the baseline value of the respective profile was within 1% to the reference baseline. As expected, the measurements with ionization chambers of 100 mm and 250 mm offer a limited accuracy, whereas an increased integration length of >=600 mm appeared to be necessary to approximate CTDI∞ in within 1%. MC simulations

  4. Predicting Electronic Failure from Smoke

    SciTech Connect

    Tanaka, T.J.

    1999-01-15

    Smoke can cause electronic equipment to fail through increased leakage currents and shorts. Sandia National Laboratories is studying the increased leakage currents caused by smoke with varying characteristics. The objective is to develop models to predict the failure of electronic equipment exposed to smoke. This requires the collection of data on the conductivity of smoke and knowledge of critical electrical systems that control high-consequence operations. We have found that conductivity is a function of the type of fuel, how it is burned, and smoke density. Video recordings of highly biased dc circuits exposed in a test chamber show that during a fire, smoke is attracted to high voltages and can build fragile carbon bridges that conduct leakage currents. The movement of air breaks the bridges, so the conductivity decreases after the fire is extinguished and the test chamber is vented. During the fire, however, electronic equipment may not operate correctly, leading to problems for critical operations dependent on electronic control. The potential for electronic failure is highly dependent on the type of electrical circuit, and Sandia National Laboratories plans to include electrical circuit modeling in the failure models.

  5. Measurement Of Gas Electron Multiplier (GEM) Detector Characteristics

    NASA Astrophysics Data System (ADS)

    Park, Seongtae; Baldelomar, Edwin; Park, Kwangjune; Sosebee, Mark; White, Andy; Yu, Jaehoon

    2011-06-01

    The High Energy Physics group of the University of Texas at Arlington has been developing gas electron multiplier detectors to use them as sensitive gap detectors in digital hadron calorimeters for the International Linear Collider, a future high energy particle accelerator. For this purpose, we constructed numerous GEM detectors that employ double GEM layers. In this study, two kinds of prototype GEM detectors were tested; one with 28×28 cm2 active area double GEM structure with a 3 mm drift gap, a 1 mm transfer gap and a 1 mm induction gap and the other with two 3×3 cm2 GEM foils in the amplifier stage with a 5 mm drift gap, a 2 mm transfer gap and a 1 mm induction gap. The detectors' characteristics from exposure to high-energy charged particles and other radiations were measured using cosmic rays and 55Fe radioactive source. From the 55Fe tests, we observed two well separated characteristic X-ray emission peaks and confirmed the detectors' functionality. We also measured chamber gains to be over 6000 at a high voltage of 395 V across each GEM electrode. The responses to cosmic rays show the spectra that fit well to Landau distributions as expected from minimum ionizing particles.

  6. Properties of a commercial PTW-60019 synthetic diamond detector for the dosimetry of small radiotherapy beams.

    PubMed

    Lárraga-Gutiérrez, José Manuel; Ballesteros-Zebadúa, Paola; Rodríguez-Ponce, Miguel; García-Garduño, Olivia Amanda; de la Cruz, Olga Olinca Galván

    2015-01-21

    A CVD based radiation detector has recently become commercially available from the manufacturer PTW-Freiburg (Germany). This detector has a sensitive volume of 0.004 mm(3), a nominal sensitivity of 1 nC Gy(-1) and operates at 0 V. Unlike natural diamond based detectors, the CVD diamond detector reports a low dose rate dependence. The dosimetric properties investigated in this work were dose rate, angular dependence and detector sensitivity and linearity. Also, percentage depth dose, off-axis dose profiles and total scatter ratios were measured and compared against equivalent measurements performed with a stereotactic diode. A Monte Carlo simulation was carried out to estimate the CVD small beam correction factors for a 6 MV photon beam. The small beam correction factors were compared with those obtained from stereotactic diode and ionization chambers in the same irradiation conditions The experimental measurements were performed in 6 and 15 MV photon beams with the following square field sizes: 10 × 10, 5 × 5, 4 × 4, 3 × 3, 2 × 2, 1.5 × 1.5, 1 × 1 and 0.5 × 0.5 cm. The CVD detector showed an excellent signal stability (<0.2%) and linearity, negligible dose rate dependence (<0.2%) and lower response angular dependence. The percentage depth dose and off-axis dose profiles measurements were comparable (within 1%) to the measurements performed with ionization chamber and diode in both conventional and small radiotherapy beams. For the 0.5 × 0.5 cm, the measurements performed with the CVD detector showed a partial volume effect for all the dosimetric quantities measured. The Monte Carlo simulation showed that the small beam correction factors were close to unity (within 1.0%) for field sizes ≥1 cm. The synthetic diamond detector had high linearity, low angular and negligible dose rate dependence, and its response was energy independent within 1% for field sizes from 1.0 to 5.0 cm. This work provides new data showing the

  7. Properties of a commercial PTW-60019 synthetic diamond detector for the dosimetry of small radiotherapy beams

    NASA Astrophysics Data System (ADS)

    Lárraga-Gutiérrez, José Manuel; Ballesteros-Zebadúa, Paola; Rodríguez-Ponce, Miguel; García-Garduño, Olivia Amanda; Olinca Galván de la Cruz, Olga

    2015-01-01

    A CVD based radiation detector has recently become commercially available from the manufacturer PTW-Freiburg (Germany). This detector has a sensitive volume of 0.004 mm3, a nominal sensitivity of 1 nC Gy-1 and operates at 0 V. Unlike natural diamond based detectors, the CVD diamond detector reports a low dose rate dependence. The dosimetric properties investigated in this work were dose rate, angular dependence and detector sensitivity and linearity. Also, percentage depth dose, off-axis dose profiles and total scatter ratios were measured and compared against equivalent measurements performed with a stereotactic diode. A Monte Carlo simulation was carried out to estimate the CVD small beam correction factors for a 6 MV photon beam. The small beam correction factors were compared with those obtained from stereotactic diode and ionization chambers in the same irradiation conditions The experimental measurements were performed in 6 and 15 MV photon beams with the following square field sizes: 10 × 10, 5 × 5, 4 × 4, 3 × 3, 2 × 2, 1.5 × 1.5, 1 × 1 and 0.5 × 0.5 cm. The CVD detector showed an excellent signal stability (<0.2%) and linearity, negligible dose rate dependence (<0.2%) and lower response angular dependence. The percentage depth dose and off-axis dose profiles measurements were comparable (within 1%) to the measurements performed with ionization chamber and diode in both conventional and small radiotherapy beams. For the 0.5 × 0.5 cm, the measurements performed with the CVD detector showed a partial volume effect for all the dosimetric quantities measured. The Monte Carlo simulation showed that the small beam correction factors were close to unity (within 1.0%) for field sizes ≥1 cm. The synthetic diamond detector had high linearity, low angular and negligible dose rate dependence, and its response was energy independent within 1% for field sizes from 1.0 to 5.0 cm. This work provides new data showing the performance of the CVD

  8. Dose rate dependence of the PTW 60019 microDiamond detector in high dose-per-pulse pulsed beams

    NASA Astrophysics Data System (ADS)

    Brualla-González, Luis; Gómez, Faustino; Pombar, Miguel; Pardo-Montero, Juan

    2016-01-01

    Recombination effects can affect the detectors used for the dosimetry of radiotherapy fields. They are important when using ionization chambers, especially in liquid-filled ionization chambers, and should be corrected for. The introduction of flattening-filter-free accelerators increases the typical dose-per-pulse used in radiotherapy beams, which leads to more important recombination effects. Diamond detectors provide a good solution for the dosimetry and quality assurance of small radiotherapy fields, due to their low energy dependence and small volume. The group of Università di Roma Tor Vergata has developed a synthetic diamond detector, which is commercialized by PTW as microDiamond detector type 60019. In this work we present an experimental characterization of the collection efficiency of the microDiamond detector, focusing on high dose-per-pulse FFF beams. The collection efficiency decreases with dose-per-pulse, down to 0.978 at 2.2 mGy/pulse, following a Fowler-Attix-like curve. On the other hand, we have found no significant dependence of the collection efficiency on the pulse repetition frequency (or pulse period).

  9. DETECTORS AND EXPERIMENTAL METHODS: A comparison of ionizing radiation damage in CMOS devices from 60Co gamma rays, electrons and protons

    NASA Astrophysics Data System (ADS)

    He, Bao-Ping; Yao, Zhi-Bin; Zhang, Feng-Qi

    2009-06-01

    Radiation hardened CC4007RH and non-radiation hardened CC4011 devices were irradiated using 60Co gamma rays, 1 MeV electrons and 1-9 MeV protons to compare the ionizing radiation damage of the gamma rays with the charged particles. For all devices examined, with experimental uncertainty, the radiation induced threshold voltage shifts (ΔVth) generated by 60Co gamma rays are equal to that of 1 MeV electron and 1-7 MeV proton radiation under 0 gate bias condition. Under 5 V gate bias condition, the distinction of threshold voltage shifts (ΔVth) generated by 60Co gamma rays and 1 MeV electrons irradiation are not large, and the radiation damage for protons below 9 MeV is always less than that of 60Co gamma rays. The lower energy the proton has, the less serious the radiation damage becomes.

  10. Test results from a comparative evaluation of a condensation nuclei fire detector

    NASA Technical Reports Server (NTRS)

    Bricker, R. W.

    1985-01-01

    The fire/smoke alarm response of a condensation nuclei fire detector (CNFD) was compared with photoelectric and ionization detectors. Tests were conducted in a former control room 8.5 m by 8.9 with a 2.7 m ceiling. The room had air supplied from above the ceiling and under the floor with return air exiting from ceiling grills. The environment was varied from 278 to 305 K and relative humidities from 8 to 65%. Four detection zones were located in the room. Each zone contained a sampling head for the CNDF, a photodetector, and an ionization detector so that each detector system had four opportunities to alarm during tests. The particle level in the test room was also monitored during tests with a condensation nuclei particle counter. The CNFD responded to 90% of exposures to smoldering plastic and 84% of exposures to visible fire. The photoelectric response was 43 and 12.5% respectively for the same conditions. The ionization response was 9 and 48 respectively.

  11. Cloud chamber visualization of primary cosmic rays

    SciTech Connect

    Earl, James A.

    2013-02-07

    From 1948 until 1963, cloud chambers were carried to the top of the atmosphere by balloons. From these flights, which were begun by Edward P. Ney at the University of Minnesota, came the following results: discovery of heavy cosmic ray nuclei, development of scintillation and cherenkov detectors, discovery of cosmic ray electrons, and studies of solar proton events. The history of that era is illustrated here by cloud chamber photographs of primary cosmic rays.

  12. Right/left assignment in drift chambers and proportional multiwire chambers (PWC's) using induced signals

    DOEpatents

    Walenta, Albert H.

    1979-01-01

    Improved multiwire chamber having means for resolving the left/right ambiguity in the location of an ionizing event. The chamber includes a plurality of spaced parallel anode wires positioned between spaced planar cathodes. Associated with each of the anode wires are a pair of localizing wires, one positioned on either side of the anode wire. The localizing wires are connected to a differential amplifier whose output polarity is determined by whether the ionizing event occurs to the right or left of the anode wire.

  13. Ionizing radiation

    Technology Transfer Automated Retrieval System (TEKTRAN)

    This chapter gives a comprehensive review on ionizing irradiation of fresh fruits and vegetables. Topics include principles of ionizing radiation, its effects on pathogenic and spoilage microorganisms, shelf-life, sensory quality, nutritional and phytochemical composition, as well as physiologic and...

  14. Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators.

    PubMed

    Tyler, Madelaine K; Liu, Paul Z Y; Lee, Christopher; McKenzie, David R; Suchowerska, Natalka

    2016-05-08

    Flattening filter-free (FFF) beams are becoming the preferred beam type for stereotactic radiosurgery (SRS) and stereotactic ablative radiation therapy (SABR), as they enable an increase in dose rate and a decrease in treatment time. This work assesses the effects of the flattening filter on small field output factors for 6 MV beams generated by both Elekta and Varian linear accelerators, and determines differences between detector response in flattened (FF) and FFF beams. Relative output factors were measured with a range of detectors (diodes, ionization cham-bers, radiochromic film, and microDiamond) and referenced to the relative output factors measured with an air core fiber optic dosimeter (FOD), a scintillation dosimeter developed at Chris O'Brien Lifehouse, Sydney. Small field correction factors were generated for both FF and FFF beams. Diode measured detector response was compared with a recently published mathematical relation to predict diode response corrections in small fields. The effect of flattening filter removal on detector response was quantified using a ratio of relative detector responses in FFF and FF fields for the same field size. The removal of the flattening filter was found to have a small but measurable effect on ionization chamber response with maximum deviations of less than ± 0.9% across all field sizes measured. Solid-state detectors showed an increased dependence on the flattening filter of up to ± 1.6%. Measured diode response was within ± 1.1% of the published mathematical relation for all fields up to 30 mm, independent of linac type and presence or absence of a flattening filter. For 6 MV beams, detector correction factors between FFF and FF beams are interchangeable for a linac between FF and FFF modes, providing that an additional uncertainty of up to ± 1.6% is accepted.

  15. Test chamber for low-background IR focal plane testing

    NASA Technical Reports Server (NTRS)

    Staller, Craig; Capps, Richard W.; Butler, Douglas; Moss, Nancy; Norwood, Wynn

    1989-01-01

    A unique and versatile vacuum chamber has been designed for JPL's IR Focal Plane Technology Group. This chamber is equipped with multiple ports for cryogen and electrical vacuum feedthroughs, pumping units, vacuum gages, sources, and detector camera heads. The design incorporates a liquid-nitrogen-cooled optical table and radiation shield for low-background IR detector testing. Focal planes can be tested at temperatures ranging from 300 K to that of liquid helium. This paper describes the design and construction of this low-background IR focal plane test chamber and discusses some of its distinctive features. An analysis of the test chamber's performance is also presented.

  16. Portable Hyperbaric Chamber

    NASA Technical Reports Server (NTRS)

    Schneider, William C. (Inventor); Locke, James P. (Inventor); DeLaFuente, Horacio (Inventor)

    2001-01-01

    A portable, collapsible hyperbaric chamber was developed. A toroidal inflatable skeleton provides initial structural support for the chamber, allowing the attendant and/or patient to enter the chamber. Oval hatches mate against bulkhead rings, and the hyperbaric chamber is pressurized. The hatches seal against an o-ring, and the internal pressure of the chamber provides the required pressure against the hatch to maintain an airtight seal. In the preferred embodiment, the hyperbaric chamber has an airlock to allow the attendant to enter and exit the patient chamber during treatment. Visual communication is provided through portholes in the patient and/or airlock chamber. Life monitoring and support systems are in communication with the interior of the hyperbaric chamber and/or airlock chamber through conduits and/or sealed feed-through connectors into the hyperbaric chamber.

  17. An assay for identification and determination of toxic rodenticide valone in serum by ion chromatography-electrospray ionization tandem mass spectrometry with ion trap detector.

    PubMed

    Cai, Mei-Qiang; Dong, Xin-Yan; Chen, Xiao-Hong; Jin, Mi-Cong

    2009-04-15

    Valone has a chronic and toxic anticoagulant rodenticide that has widely used in China and has resulted in some accidental and intentional intoxication in recent years. The literature reported so far lacks sensitive and selective method for the confirmation of valone. The purpose of this study was to establish a novel assay for the identification and quantification of valone in serum by ion chromatography-electrospray ionization tandem mass spectrometry (IC-MS/MS). After serum sample was extracted with methanol/acetonitrile (10:90, v/v) and cleaned by Oasis HLB solid-phase extraction cartridge, chromatographic separation was performed on an Ionpac AS11 column with an eluent of methanol/30 mmol/L KOH (10:90, v/v). The overall extraction efficiency was >81.0%, and the limit of quantification was 0.5 ng/mL for valone. Regression analysis of the calibration data revealed good correlation (r(2)>0.99) for valone. Intra- and inter-day precisions for quality-control samples were less than 8.0 and 13.7%, respectively. The proposed method enables the identification and quantification of valone in both clinical and forensic specimens.

  18. High efficiency photoionization detector

    DOEpatents

    Anderson, D.F.

    1984-01-31

    A high efficiency photoionization detector is described using tetraaminoethylenes in a gaseous state having a low ionization potential and a relative photoionization cross section which closely matches the emission spectrum of xenon gas. Imaging proportional counters are also disclosed using the novel photoionization detector of the invention. The compound of greatest interest is TMAE which comprises tetrakis(dimethylamino)ethylene which has a measured ionization potential of 5.36 [+-] 0.02 eV, and a vapor pressure of 0.35 torr at 20 C. 6 figs.

  19. High efficiency photoionization detector

    DOEpatents

    Anderson, David F.

    1984-01-01

    A high efficiency photoionization detector using tetraaminoethylenes in a gaseous state having a low ionization potential and a relative photoionization cross section which closely matches the emission spectrum of xenon gas. Imaging proportional counters are also disclosed using the novel photoionization detector of the invention. The compound of greatest interest is TMAE which comprises tetrakis(dimethylamino)ethylene which has a measured ionization potential of 5.36.+-.0.02 eV, and a vapor pressure of 0.35 torr at 20.degree. C.

  20. MIMAC-He3: MICRO-TPC MATRIX OF CHAMBERS OF 3He

    NASA Astrophysics Data System (ADS)

    Santos, D.; Guillaudin, O.; Lamy, Th.; Mayet, F.; Moulin, E.

    2007-08-01

    The project of a micro-TPC matrix of chambers of 3He for direct detection of non-baryonic dark matter is outlined. The privileged properties of 3He are highlighted. The double detection (ionization - projection of tracks) will assure the electron-recoil discrimination. The complementarity of MIMAC-He3 for supersymmetric dark matter search with respect to other experiments is illustrated. The modular character of the detector allows to have different gases to get A-dependence. The pressure degreee of freedom gives the possibility to work at high and low pressure. The low pressure regime gives the possibility to get the directionality of the tracks. The first measurements of ionization at very few keVs for 3He in 4He gas are described.