Sample records for joint degree program

  1. Joint MS Degree Program between the Korea University of Technology and Education and the University of Kansas

    NASA Astrophysics Data System (ADS)

    Dougherty, R. L.; Kim, Kwang Su

    This paper provides an overview of the Joint MS Degree Program between Korea University of Technology and Education’ s (KUT) Mechatronics Department and Kansas University’ s (KU) Mechanical Engineering Department. Discussions were initiated in early 2005 which resulted in a formal agreement being approved by both parties in mid-2007. The Joint MS Degree Program is composed of 30 semester credit hours, equally split between the two universities, with the actual degree being awarded by the institution at which the thesis work is performed. Issues addressed during the development of this Joint MS Program included: joint versus dual degrees, institutional acceptance of the transfer of fifteen hours of credit for an MS degree, different admissions requirements and procedures for the two institutions, financial support of the students, faculty advisors at each institution, Graduate Directors at each institution, transcript acknowledgement of the Joint Degree, residency requirements, English speaking requirements/abilities, thesis publication allowances/requirements, and time zone differences for virtual meetings. These issues have been addressed, and the Joint MS Degree Program is functioning with a small number of students having taken advantage of the opportunity since the Program’ s inception. Future considerations are: growing the number of students in the Program, expansion to other Departments besides KU-Mechanical Engineering and KUTMechatronics, including other universities in the Program, expansion to a Joint PhD Degree Program, and stronger funding resources.

  2. Quality Assurance of Joint Degree Programs from the Perspective of Quality Assurance Agencies: Experience in East Asia

    ERIC Educational Resources Information Center

    Hou, Yung-Chi; Ince, Martin; Tsai, Sandy; Wang, Wayne; Hung, Vicky; Lin Jiang, Chung; Chen, Karen Hui-Jung

    2016-01-01

    Joint degree programs have gained popularity in East Asia, due to the growth of transnational higher education in the region since 2000. However, the external quality assurance (QA) and accreditation of joint degree programs is a challenge for QA agencies, as it normally involves the engagement of several institutions and multiple national…

  3. Strategies of Supporting Chinese Students in an International Joint Degree Program

    ERIC Educational Resources Information Center

    Arshakian, Arakssi; Wang, Vivian

    2017-01-01

    The international joint degree program is one of the recent ways of international collaborations in Higher Education. Those programs involve intensive academic collaborations as well as institutional alliance.?Such programs could provide a supportive environment for international students through international partnerships. The article provides a…

  4. Dual Degree Programs at the University of Kentucky College of Pharmacy

    PubMed Central

    Senft, Sandra L.; Thompson, Chad

    2008-01-01

    The rapid growth and evolution of the pharmacy profession has created a wide array of opportunities for graduating pharmacists beyond traditional community pharmacy or hospital practice. Management and leadership positions in federal and state healthcare agencies, pharmaceutical companies, hospitals, retail pharmacies, academia and managed care organizations increasingly require the pharmaceutical knowledge obtained through a doctor of pharmacy (PharmD) degree combined with financial, organizational, and management skills. In these innovative positions, pharmacists are being called upon to assume responsibilities as executives and administrators in systems providing pharmacist care services to patients. To endow students with knowledge and skills required to perform the duties required in these decision-making positions, the University of Kentucky College of Pharmacy has established 3 joint degree programs: the PharmD/Master of Business Administration (PharmD/MBA), PharmD/Master of Public Administration (PharmD/MPA), and PharmD/Master of Science in Economics (PharmD/MS). This paper describes these joint degree programs. PMID:18322574

  5. Normal axial alignment of the lower extremity and load-bearing distribution at the knee.

    PubMed

    Hsu, R W; Himeno, S; Coventry, M B; Chao, E Y

    1990-06-01

    Based on a series of 120 normal subjects of different gender and age, the geometry of the knee joint was analyzed using a full-length weight-bearing roentgenogram of the lower extremity. A special computer program based on the theory of a rigid body spring model was applied to calculate the important anatomic and biomechanical factors of the knee joint. The tibiofemoral mechanical angle was 1.2 degrees varus. Hence, it is difficult to rationalize the 3 degree varus placement of the tibial component in total knee arthroplasty suggested by some authors. The distal femoral anatomic valgus (measured from the lower one-half of the femur) was 4.2 degrees in reference to its mechanical axis. This angle became 4.9 degrees when the full-length femoral anatomic axis was used. When simulating a one-legged weight-bearing stance by shifting the upper-body gravity closer to the knee joint, 75% of the knee joint load passed through the medial tibial plateau. The knee joint-line obliquity was more varus in male subjects. The female subjects had a higher peak joint pressure and a greater patello-tibial Q angle. Age had little effect on the factors relating to axial alignment of the lower extremity and load transmission through the knee joint.

  6. Aeroacoustics Research Program in JIAFS

    NASA Technical Reports Server (NTRS)

    Myers, Michael K.

    2000-01-01

    This paper presents a final report on Aeroacoustics Research Program in JIAFS (Joint Institute For Advancement of Flight Sciences). The objectives of the program were to conduct research at the NASA Langley Research Center and to provide a comprehensive education program at the Center leading to advanced degrees in aeroacoustics.

  7. Joint Supervision of Research Degrees: Second Thoughts.

    ERIC Educational Resources Information Center

    Bourner, Tom; Hughes, Mark

    1991-01-01

    A discussion of Britain's regulation that master's and doctoral degree candidates must have two or three program supervisors identifies four potential problems: fragmentation of supervisory responsibilities; conflicting advice; unproductive games; and absence of overall perspective on the thesis. Experience with an improved, team approach is…

  8. [New ways of higher education in nursing: globalisation of nursing leadership and its teaching--dual degree in nursing].

    PubMed

    Pop, Marcel; Hollós, Sándor; Vingender, István; Mészáros, Judit

    2009-03-08

    Our paper is presenting a new initiative regarding an international cooperation willing to develop a dual degree program in nursing, the so-called Transatlantic Curriculum in Nursing. The candidates--after successful completion of their studies--will get a European and an American partner diploma in nursing. The objective is to prepare an internationally and culturally competent workforce; develop the practice of nursing students' exchange programs; process the model of dual degree independent of geographical, political or cultural borders; spread the evidence-based nursing standards in the daily practice. The partners in this initiative are Semmelweis University in Budapest, Hungary, Nazareth College of Rochester, NY, USA and Laurea University in Tikkurila, Finland. The planned activities in the framework of the program: mutual student and staff mobility, joint curriculum development and teaching process, determining joint standards. The expected outcomes are: to develop a standardised model for the enhancement and implementation of international educational programs in nursing; to improve institutional work culture; to improve professional terminology and cultural abilities; to create the model of a new type of nursing professional having a high level of cultural and language competence which are indispensable for participating in global programs.

  9. Community Colleges and Higher Education: A Joint Venture.

    ERIC Educational Resources Information Center

    Chambers, Sharon M.; Weeks, Sandy R.; Chaloupka, Larry J.

    2003-01-01

    Describes the teacher preparation program at Colin County Community College District (CCCCD), Texas, the first community college program in the nation authorized to prepare teachers. Also reports that Texas A&M University-Commerce has joined with CCCCD to provide a bridge between the community college program and graduate degrees at the…

  10. Innovating in Health Care Management Education: Development of an Accelerated MBA and MPH Degree Program at Yale

    PubMed Central

    Forman, Howard P.; Pistell, Anne F.; Nembhard, Ingrid M.

    2015-01-01

    Increasingly, there is recognition of the need for individuals with expertise in both management and public health to help health care organizations deliver high-quality and cost-effective care. The Yale School of Public Health and Yale School of Management began offering an accelerated Master of Business Administration (MBA) and Master of Public Health (MPH) joint degree program in the summer of 2014. This new program enables students to earn MBA and MPH degrees simultaneously from 2 fully accredited schools in 22 months. Students will graduate with the knowledge and skills needed to become innovative leaders of health care organizations. We discuss the rationale for the program, the developmental process, the curriculum, benefits of the program, and potential challenges. PMID:25706023

  11. Innovating in health care management education: development of an accelerated MBA and MPH degree program at Yale.

    PubMed

    Pettigrew, Melinda M; Forman, Howard P; Pistell, Anne F; Nembhard, Ingrid M

    2015-03-01

    Increasingly, there is recognition of the need for individuals with expertise in both management and public health to help health care organizations deliver high-quality and cost-effective care. The Yale School of Public Health and Yale School of Management began offering an accelerated Master of Business Administration (MBA) and Master of Public Health (MPH) joint degree program in the summer of 2014. This new program enables students to earn MBA and MPH degrees simultaneously from 2 fully accredited schools in 22 months. Students will graduate with the knowledge and skills needed to become innovative leaders of health care organizations. We discuss the rationale for the program, the developmental process, the curriculum, benefits of the program, and potential challenges.

  12. A Survey of Collaborative Projects of Maryland Postsecondary Institutions.

    ERIC Educational Resources Information Center

    Maryland State Higher Education Commission, Annapolis.

    This report, prepared at the request of an informal Maryland higher education workgroup called "A United Voice for Higher Education," is a summary of various campus collaborative activities. Activities were grouped in the following categories: academic, including joint degree programs, articulation programs, and distance learning…

  13. To What Extent Is Criminal Justice Content Specifically Addressed in MSW Programs?

    ERIC Educational Resources Information Center

    Epperson, Matthew W.; Roberts, Leslie E.; Ivanoff, Andre; Tripodi, Stephen J.; Gilmer, Christy N.

    2013-01-01

    This study examined the extent to which criminal justice content is addressed in all CSWE-accredited MSW programs in the United States ("N"?=?192). Criminal justice content was measured in three areas: (1) dual or joint degree programs, (2) concentrations or specializations, and (3) coursework. Excluding social work and law classes, 22%…

  14. Associate Degree Nursing: Model Prerequisites Validation Study. California Community College Associate Degree Programs by The Center for Student Success, A Health Care Initiative Sponsored Project.

    ERIC Educational Resources Information Center

    Phillips, Brad C.; Spurling, Steven; Armstrong, William A.

    California faces a severe nursing shortage, with the number of registered nurses far below what is required to avert a potential state health care crisis. The Associate Degree Nursing (ADN) Project is a joint project involving scholars, educational researchers, and analysts from the Center for Student Success (CSS) housed at City College of San…

  15. Cultural Influences on the Design and Management of Transnational Higher Education Programs in China: A Case Study of Three Programs

    ERIC Educational Resources Information Center

    He, Lan; Liu, Ersi

    2018-01-01

    Purpose: The purpose of this paper is to identify the influence of cultural differences on the design and management of two franchise programs and one joint degree program run by a Chinese university in partnership with a British university and a French university. The cross-cultural challenges and the differences in the strategies taken by the…

  16. The inception and evolution of a unique masters program in cancer biology, prevention and control.

    PubMed

    Cousin, Carolyn; Blancato, Jan

    2010-09-01

    The University of the District of Columbia (UDC) and the Lombardi Comprehensive Cancer Center (LCCC), Georgetown University Medical Center established a Masters Degree Program in Cancer Biology, Prevention and Control at UDC that is jointly administered and taught by UDC and LCCC faculty. The goal of the Masters Degree Program is to educate students as master-level cancer professionals capable of conducting research and service in cancer biology, prevention, and control or to further advance the education of students to pursue doctoral studies. The Program's unique nature is reflected in its philosophy "the best cancer prevention and control researchers are those with a sound understanding of cancer biology". This program is a full-time, 2-year, 36-credit degree in which students take half of their coursework at UDC and half of their coursework at LCCC. During the second year, students are required to conduct research either at LCCC or UDC. Unlike most cancer biology programs, this unique Program emphasizes both cancer biology and cancer outreach training.

  17. Development of an Interdisciplinary, Intercultural Master's Program on Sustainability: Learning from the Richness of Diversity

    ERIC Educational Resources Information Center

    van Dam-Mieras, Rietje; Lansu, Angelique; Rieckmann, Marco; Michelsen, Gerd

    2008-01-01

    The purpose of this article is to describe a joint effort between three European and six Latin American universities to create an international Master's degree program on Sustainable Development and Management. Faculty members from these institutions are working together on this unusual and innovative project, which recognizes the importance of…

  18. Team-Based Programs: The EMP. An Industry/University Partnership-The Educational Model for the 21st Century.

    ERIC Educational Resources Information Center

    Haynes, Ray; And Others

    California Polytechnic State University's College of Business and College of Engineering have joined forces to create a joint Engineering Management Program (EMP). Students holding undergraduate engineering or equivalent degrees enter and earn both Masters in Business Administration and Masters of Science in Engineering in 24 months. The program…

  19. The RITE Approach to Agile Acquisition

    DTIC Science & Technology

    2013-04-01

    SSCPAC. He currently provides project management and technical support to the Global Command and Control System–Joint (GCCS-J) Integrated Imagery and...the Strike Planning and Execution Systems Program Office (PMA-281). Boyce has a Bachelor of Science degree in information systems management and is...has since provided dedicated professional service to the USAF, NAVSUP, and SSCPAC. Roussel has a Bachelor of Science degree in management and a Master

  20. Software for Simulating a Complex Robot

    NASA Technical Reports Server (NTRS)

    Goza, S. Michael

    2003-01-01

    RoboSim (Robot Simulation) is a computer program that simulates the poses and motions of the Robonaut a developmental anthropomorphic robot that has a complex system of joints with 43 degrees of freedom and multiple modes of operation and control. RoboSim performs a full kinematic simulation of all degrees of freedom. It also includes interface components that duplicate the functionality of the real Robonaut interface with control software and human operators. Basically, users see no difference between the real Robonaut and the simulation. Consequently, new control algorithms can be tested by computational simulation, without risk to the Robonaut hardware, and without using excessive Robonaut-hardware experimental time, which is always at a premium. Previously developed software incorporated into RoboSim includes Enigma (for graphical displays), OSCAR (for kinematical computations), and NDDS (for communication between the Robonaut and external software). In addition, RoboSim incorporates unique inverse-kinematical algorithms for chains of joints that have fewer than six degrees of freedom (e.g., finger joints). In comparison with the algorithms of OSCAR, these algorithms are more readily adaptable and provide better results when using equivalent sets of data.

  1. Determination of consistent patterns of range of motion in the ankle joint with a computed tomography stress-test.

    PubMed

    Tuijthof, Gabriëlle Josephine Maria; Zengerink, Maartje; Beimers, Lijkele; Jonges, Remmet; Maas, Mario; van Dijk, Cornelis Niek; Blankevoort, Leendert

    2009-07-01

    Measuring the range of motion of the ankle joint can assist in accurate diagnosis of ankle laxity. A computed tomography-based stress-test (3D CT stress-test) was used that determines the three-dimensional position and orientation of tibial, calcaneal and talar bones. The goal was to establish a quantitative database of the normal ranges of motion of the talocrural and subtalar joints. A clinical case on suspected subtalar instability demonstrated the relevance the proposed method. The range of motion was measured for the ankle joints in vivo for 20 subjects using the 3D CT stress-test. Motion of the tibia and calcaneus relative to the talus for eight extreme foot positions were described by helical parameters. High consistency for finite helical axis orientation (n) and rotation (theta) was shown for: talocrural extreme dorsiflexion to extreme plantarflexion (root mean square direction deviation (eta) 5.3 degrees and theta: SD 11.0 degrees), talorucral and subtalar extreme combined eversion-dorsiflexion to combined inversion-plantarflexion (eta: 6.7 degrees , theta: SD 9.0 degrees and eta:6.3 degrees , theta: SD 5.1 degrees), and subtalar extreme inversion to extreme eversion (eta: 6.4 degrees, theta: SD 5.9 degrees). Nearly all dorsi--and plantarflexion occurs in the talocrural joint (theta: mean 63.3 degrees (SD 11 degrees)). The inversion and internal rotation components for extreme eversion to inversion were approximately three times larger for the subtalar joint (theta: mean 22.9 degrees and 29.1 degrees) than for the talocrural joint (theta: mean 8.8 degrees and 10.7 degrees). Comparison of the ranges of motion of the pathologic ankle joint with the healthy subjects showed an increased inversion and axial rotation in the talocrural joint instead of in the suspected subtalar joint. The proposed diagnostic technique and the acquired database of helical parameters of ankle joint ranges of motion are suitable to apply in clinical cases.

  2. Determining Metacarpophalangeal Flexion Angle Tolerance for Reliable Volumetric Joint Space Measurements by High-resolution Peripheral Quantitative Computed Tomography.

    PubMed

    Tom, Stephanie; Frayne, Mark; Manske, Sarah L; Burghardt, Andrew J; Stok, Kathryn S; Boyd, Steven K; Barnabe, Cheryl

    2016-10-01

    The position-dependence of a method to measure the joint space of metacarpophalangeal (MCP) joints using high-resolution peripheral quantitative computed tomography (HR-pQCT) was studied. Cadaveric MCP were imaged at 7 flexion angles between 0 and 30 degrees. The variability in reproducibility for mean, minimum, and maximum joint space widths and volume measurements was calculated for increasing degrees of flexion. Root mean square coefficient of variance values were < 5% under 20 degrees of flexion for mean, maximum, and volumetric joint spaces. Values for minimum joint space width were optimized under 10 degrees of flexion. MCP joint space measurements should be acquired at < 10 degrees of flexion in longitudinal studies.

  3. Celebrating minority professionals in forestry and natural resources conservation: proceedings of the symposium on the tenth anniversary of the 2 + 2 joint degree program in forestry and natural resources conservation

    Treesearch

    Oghenekome U. Onokpise; Don L. Rockwood; Dreamal H. Worthen; Ted Willis

    2008-01-01

    The 22 papers in this symposium highlight the program and its contribution to increasing minority professionals in forestry and natural resources conservation. The tenth anniversary symposium brought together graduates of the program, current students and officials from the universities, the U.S. Forest Service, other agencies, and private industry. The theme of the...

  4. Kinematically redundant robot manipulators

    NASA Technical Reports Server (NTRS)

    Baillieul, J.; Hollerbach, J.; Brockett, R.; Martin, D.; Percy, R.; Thomas, R.

    1987-01-01

    Research on control, design and programming of kinematically redundant robot manipulators (KRRM) is discussed. These are devices in which there are more joint space degrees of freedom than are required to achieve every position and orientation of the end-effector necessary for a given task in a given workspace. The technological developments described here deal with: kinematic programming techniques for automatically generating joint-space trajectories to execute prescribed tasks; control of redundant manipulators to optimize dynamic criteria (e.g., applications of forces and moments at the end-effector that optimally distribute the loading of actuators); and design of KRRMs to optimize functionality in congested work environments or to achieve other goals unattainable with non-redundant manipulators. Kinematic programming techniques are discussed, which show that some pseudo-inverse techniques that have been proposed for redundant manipulator control fail to achieve the goals of avoiding kinematic singularities and also generating closed joint-space paths corresponding to close paths of the end effector in the workspace. The extended Jacobian is proposed as an alternative to pseudo-inverse techniques.

  5. Transfer and Joint Programs - do they Work?

    NASA Astrophysics Data System (ADS)

    Chapman, P.; Chang, P.; Wu, D.

    2013-12-01

    With the increased interest in internationalization, many western universities have been working to attract students from China and other Asian countries. Texas A&M University has had a collaborative Ph.D. program with Ocean University, Qingdao, China since 2008. Chinese students spend their first one or two years in Qingdao, then transfer to TAMU to complete their coursework and carry out research. The program has so far produced 6 PhDs and one MS degree, and another 14 students are registered through the program. We anticipate sending U.S. students to China on a reciprocal basis, although the lack of Chinese-speaking U.S. students is an issue. Additionally, the Memorandum of Understanding has led to joint research projects between the two universities, with publications in top journals. Novel aspects of the program include joint funding by the U.S. and China, co-chairing graduate committees by both university faculties, interviewing students in Qingdao before they are accepted by TAMU, and initial studies in one country and then transfer to the other. Such programs require 'heroes' on each side to set up and continue the program who trust each other, as well as support from the upper administration. Even with such support, outside influences and different cultures can affect the effectiveness of the program.

  6. Samling pa mattan: En utvardering av integrerad skolbarnsomsorg (SKOBO) inom Finspangs Kommun (Integration of Pre-School and School Children).

    ERIC Educational Resources Information Center

    Bergeling, Ann-Sofie; Gustavsson, Maria

    This study evaluated a project for integrating preschool- and school-age children in a joint educational program. The integration project (SKOBO) has been implemented in all education units in the investigated municipality. Each school has, however, had a considerable degree of freedom in carrying out the program with the potential to find its…

  7. Hip Surveillance in Children with Cerebral Palsy.

    PubMed

    Huser, Aaron; Mo, Michelle; Hosseinzadeh, Pooya

    2018-04-01

    The hip is the second most common involved joint in cerebral palsy. Hip displacement occurs in more than 33% of children with cerebral palsy, with a higher prevalence in nonambulatory children. Hip displacement in this population is typically progressive. Hip dislocation can result in pain and difficulty with sitting and perineal care. Since early stage of hip displacement can be silent, and hip surveillance programs are recommended. Most programs use the degree of hip dysplasia and Growth Motor Function Classification System level for screening recommendations. Treatment depends on the degree of dysplasia, functional status of the patient, and patient's age. Copyright © 2017 Elsevier Inc. All rights reserved.

  8. Dynamic Modelling Of A SCARA Robot

    NASA Astrophysics Data System (ADS)

    Turiel, J. Perez; Calleja, R. Grossi; Diez, V. Gutierrez

    1987-10-01

    This paper describes a method for modelling industrial robots that considers dynamic approach to manipulation systems motion generation, obtaining the complete dynamic model for the mechanic part of the robot and taking into account the dynamic effect of actuators acting at the joints. For a four degree of freedom SCARA robot we obtain the dynamic model for the basic (minimal) configuration, that is, the three degrees of freedom that allow us to place the robot end effector in a desired point, using the Lagrange Method to obtain the dynamic equations in matrix form. The manipulator is considered to be a set of rigid bodies inter-connected by joints in the form of simple kinematic pairs. Then, the state space model is obtained for the actuators that move the robot joints, uniting the models of the single actuators, that is, two DC permanent magnet servomotors and an electrohydraulic actuator. Finally, using a computer simulation program written in FORTRAN language, we can compute the matrices of the complete model.

  9. Coactivation of the elbow antagonist muscles is not affected by the speed of movement in isokinetic exercise.

    PubMed

    Bazzucchi, Ilenia; Sbriccoli, Paola; Marzattinocci, Giulia; Felici, Francesco

    2006-02-01

    Since muscle coactivation increases the stiffness and stability of a joint, greater coactivation is likely during faster than slower movements. Very few studies, though, have been conducted to verify this hypothesis. Moreover, a large number of studies have examined coactivation of muscles surrounding the knee joint whereas there are few reports on the elbow joint. The aim of this study was therefore to compare the antagonist activation of the elbow flexors and extensors during isokinetic concentric exercises and to investigate the influence of angular velocity on their activation. Twelve men participated in the study. The surface electromyographic signals (sEMG) were recorded from the biceps brachii (BB) and triceps brachii (TB) muscles during three maximal voluntary isometric contractions (MVC) of elbow flexors and extensors and a set of three maximal elbow flexions and extensions each at 15 degrees, 30 degrees , 60 degrees, 120 degrees, 180 degrees, and 240 degrees.s(-1). Normalized root mean square (RMS) of sEMG was calculated during the isokinetic phase of movement as an index of sEMG amplitude. During elbow flexion, the antagonist activation of BB averaged 16.2% lower than TB, and this difference was statistically significant at all angular velocities. The normalized RMS values ranged from 26.0% +/- 19.0 at MVC to 37.8% +/- 13.9 at 240 degrees.s(-1) for antagonist TB activation, and from 5.7% +/- 5.2 at MVC to 18.9% +/- 8.6 at 240 degrees.s(-1) for antagonist BB activation. No influence of angular velocity on agonist and antagonist activity was found. Moreover, flexion and extension torques were both strongly affected by the amount of antagonist activation. The functional specialization of the two muscle groups could be responsible for the different levels of antagonist activation. The frequent use of BB, which is not assisted by gravity during daily activities, could lead to reduced coactivation due to a better functioning of the control system based upon reciprocal innervation. These findings may have significant implications in the design of rehabilitation programs directed to the elbow joint.

  10. 42 CFR 486.104 - Condition for coverage: Qualifications, orientation and health of technical personnel.

    Code of Federal Regulations, 2010 CFR

    2010-10-01

    ...), or (4) of this section: (1) Successful completion of a program of formal training in X-ray technology in a school approved by the Joint Review Committee on Education in Radiologic Technology (JRCERT), or have earned a bachelor's or associate degree in radiologic technology from an accredited college or...

  11. Utilizing Video Conferencing to Introduce an International Perspective to Foundation Practice

    ERIC Educational Resources Information Center

    Forgey, Mary Ann; Loughran, Hilda; Hansen, Johna

    2013-01-01

    Video conferencing has much potential to enrich international social work education. In this educational initiative, video conferencing was used to deliver a joint foundation social work practice class to students attending an MSW degree program both in the United States and in Ireland. Student feedback indicated that they gained an appreciation…

  12. Communication Instruction in a Mature Institutional Partnership: An Examination of Evolving Methods

    ERIC Educational Resources Information Center

    Thomas, Martha Wetterhall; Hardy, Samuel B., IV

    2005-01-01

    Since 1994, the University of South Carolina at Columbia and the Wirtschaftsuniversitat Wien (Vienna University of Economics and Business Administration) have offered a joint master's degree in international business. Communication instruction was initially a stable component of the program, with a week-long course at the beginning and a…

  13. Sedgemoor: A Suitable Case for Treatment? Heritage, Interpretation and Educational Process.

    ERIC Educational Resources Information Center

    Thompson, Lynne

    A partnership between the Universities of Exeter and Bournemouth at their joint University Centre in Yeovil College, Somerset (England) allowed local students to participate in higher education via a BA degree in Heritage and Regional Studies. This program represents several disciplines including history, literature, and the visual arts. It aims…

  14. Results in Developing an Engineering Degree Program in Safeguards and Security of Nuclear Materials at Moscow Engineering Physics Institute

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kryuchkov, Eduard F.; Geraskin, Nikolay I.; Killinger, Mark H.

    The world’s first master’s degree program in nuclear safeguards and security, established at Moscow Engineering Physics Institute (MEPhI), has now graduated nine classes of students. Most of the graduates have gone on to work at government agencies, research organizations, or obtain their PhD. In order to meet the demand for safeguards and security specialists at nuclear facilities, MEPhI established a 5½ year engineering degree program that provides more hands-on training desired by facilities. In February 2004, the first students began their studies in the new discipline Nuclear Material Safeguards and Nonproliferation. This class, as well as other subsequent classes, includedmore » students who started the program in their third year of studies, as the first 2½ years consists of general engineering curriculum. Fourteen students made up the first graduating class, receiving their engineering degrees in February 2007. The topics addressed in this paper include specific features of the program caused by peculiarities of Russian education legislation and government quality control of academic education. This paper summarizes the main joint actions undertaken by MEPhI and the US National Laboratories in conjunction with the U.S. Department of Energy, to develop the engineering degree program. Also discussed are the program’s specific training requirements, student internships, and job placement. The paper concludes with recommendations from a recent international seminar on nonproliferation education and training.« less

  15. Program of Research in Aeronautics

    NASA Technical Reports Server (NTRS)

    1981-01-01

    A prospectus of the educational and research opportunities available at the Joint Institute for Advancement of Flight Sciences, operated at NASA Langley Research Center in conjunction with George Washington University's School of Engineering and Applied Sciences is presented. Requirements of admission to various degree programs are given as well as the course offerings in the areas of acoustics, aeronautics, environmental modelling, materials science, and structures and dynamics. Research facilities for each field of study are described. Presentations and publications (including dissertations and theses) generated by each program are listed as well as faculty members visting scientists and engineers.

  16. Study of the Peak Shear Strength of a Cement-Filled Hard Rock Joint

    NASA Astrophysics Data System (ADS)

    She, Cheng-Xue; Sun, Fu-Ting

    2018-03-01

    The peak shear strength of a cement-filled hard rock joint is studied by theoretical analysis and laboratory testing. Based on the concept of the shear resistance angle, by combining the statistical method and fractal theory, three new parameters are proposed to characterize the three-dimensional joint morphology, reflecting the effects of the average roughness, multi-scale asperities and the dispersion degree of the roughness distribution. These factors are independent of the measurement scale, and they reflect the anisotropy of the joint roughness. Compressive shear tests are conducted on cement-filled joints. Because joints without cement can be considered special cement-filled joints in which the filling degree of cement is zero, they are also tested. The cement-filled granite joint fails primarily along the granite-cement interfaces. The filling degree of cement controls the joint failure and affects its mechanical behaviour. With a decrease in the filling degree of cement, the joint cohesion decreases; however, the dilatancy angle and the basic friction angle of the interface increase. As the filling degree approaches zero, the cohesion approaches zero, while the dilatancy angle and the basic friction angle increase to those of the joint without cement. A set of formulas is proposed to evaluate the peak shear strength of the joints with and without cement. The formulas are shown to be reasonable by comparison with the tested peak shear strength, and they reflect the anisotropy of the strength. This research deepens the understanding of cement-filled joints and provides a method to evaluate their peak shear strength.

  17. [Arthroscopic therapy of ankle joint impingement syndrome after operation of ankle joint fracture dislocation].

    PubMed

    Feng, Zhibin; Mi, Kun; Wei, Renzhi; Liu, Wu; Wang, Bin

    2011-07-01

    To study the operative procedure and the effectiveness of arthroscopic therapy for ankle joint impingement syndrome after operation of ankle joint fracture dislocation. Between March 2008 and April 2010, 38 patients with ankle joint impingement syndrome after operation of ankle joint fracture dislocation were treated. Among them, there were 28 males and 10 females with an average age of 28 years (range, 18 to 42 years). The time from internal fixation to admission was 12-16 months (mean, 13.8 months). There were pressing pain in anterolateral and anterior ankle. The dorsal extension ranged from -20 to -5 degrees (mean, -10.6 degrees), and the palmar flexion was 30-40 degrees (mean, 35.5 degrees). The total score was 48.32 +/- 9.24 and the pain score was 7.26 +/- 1.22 before operation according to American Orthopaedic Foot and Ankle Society (AOFAS) ankle and hindfoot score system. The X-ray films showed osteophyte formation in anterior tibia and talus; MRI showed cartilage injury in 22 cases. Arthroscopic intervention included removing osteophytes, debriding fabric scars and synovial membrane tissues, and removing osteochondral fragments. Arthroscopic microfracture technique was used in 22 patients with cartilage injury. All incisions healed primarily. Thirty-eight cases were followed up 10-26 months (mean, 16 months). At last follow-up, 26 patients had normal range of motion (ROM); the dorsal extension was 15-25 degrees (mean, 19.6 degrees) and the palmar flexion was 35-45 degrees (mean, 40.7 degrees). Eight patients had mild limited ROM; the dorsal extension was 5-15 degrees (mean, 7.2 degrees) and the palmar flexion was 35-45 degrees (mean, 39.5 degrees). Four patients had mild limited ROM and pain in posterior portion of the ankle after a long walking (3-4 hours); the dorsal extension was 0-5 degrees (mean, 2.6 degrees) and the palmar flexion was 35-40 degrees (mean, 37.5 degrees). The total score was 89.45 +/- 9.55 and the pain score was 1.42 +/- 1.26 after operation according to AOFAS ankle and hindfoot score system, showing significant differences when compared with preoperative ones (t=21.962, P=0.000; t=16.762, P=0.000). Arthroscopic treatment of ankle joint impingement syndrome after operation of ankle joint fracture dislocation is an effective, simple, and safe method.

  18. Flexor bias of joint position in humans during spaceflight

    NASA Technical Reports Server (NTRS)

    McCall, G. E.; Goulet, C.; Boorman, G. I.; Roy, R. R.; Edgerton, V. R.

    2003-01-01

    The ability to estimate ankle and elbow joint position was tested before, during, and after a 17-day spaceflight. Subjects estimated targeted joint angles during isovelocity (IsoV) joint movements with agonist muscle groups either active or relaxed. These movements included elbow extension (EE) and elbow flexion (EF), and plantarflexion (PF) and dorsiflexion (DF) of the ankle. Subjects also estimated these joint positions while moving the dynamometer at their chosen (variable) velocity (VarV) during EE and PF. For IsoV tests, no differences were observed between active and passive movements for either the ankle or elbow. Compared with those of pre-flight test days, estimates of targeted elbow joint angles were approximately 5 degrees to 15 degrees more flexed in-flight, and returned toward the pre-flight values during recovery. The spaceflight effects for the ankle were inconsistent and less prevalent than those for the elbow. The VarV PF test condition for the 120 degrees target angle at the ankle exhibited approximately 5 degrees to 7 degrees more DF target angle estimates in-flight compared with those pre- or post-flight. In contrast, during IsoV PF there was a tendency for ankle estimates to be approximately 2 degrees to 3 degrees more PF after 2-3 days exposure to spaceflight. These data indicate that during spaceflight the perception of elbow extension is greater than actuality, and are consistent with the interpretation that microgravity induced a flexor bias in the estimation of the actual elbow joint position. Moreover, these effects in joint proprioception during spaceflight were observed in individual isolated single-joint movements during tasks in which vestibular function in maintaining posture were minimal.

  19. Flexor bias of joint position in humans during spaceflight.

    PubMed

    McCall, G E; Goulet, C; Boorman, G I; Roy, R R; Edgerton, V R

    2003-09-01

    The ability to estimate ankle and elbow joint position was tested before, during, and after a 17-day spaceflight. Subjects estimated targeted joint angles during isovelocity (IsoV) joint movements with agonist muscle groups either active or relaxed. These movements included elbow extension (EE) and elbow flexion (EF), and plantarflexion (PF) and dorsiflexion (DF) of the ankle. Subjects also estimated these joint positions while moving the dynamometer at their chosen (variable) velocity (VarV) during EE and PF. For IsoV tests, no differences were observed between active and passive movements for either the ankle or elbow. Compared with those of pre-flight test days, estimates of targeted elbow joint angles were approximately 5 degrees to 15 degrees more flexed in-flight, and returned toward the pre-flight values during recovery. The spaceflight effects for the ankle were inconsistent and less prevalent than those for the elbow. The VarV PF test condition for the 120 degrees target angle at the ankle exhibited approximately 5 degrees to 7 degrees more DF target angle estimates in-flight compared with those pre- or post-flight. In contrast, during IsoV PF there was a tendency for ankle estimates to be approximately 2 degrees to 3 degrees more PF after 2-3 days exposure to spaceflight. These data indicate that during spaceflight the perception of elbow extension is greater than actuality, and are consistent with the interpretation that microgravity induced a flexor bias in the estimation of the actual elbow joint position. Moreover, these effects in joint proprioception during spaceflight were observed in individual isolated single-joint movements during tasks in which vestibular function in maintaining posture were minimal.

  20. A joint program with Japanese investigators to map carbon 2 line emission from the galaxy

    NASA Technical Reports Server (NTRS)

    Low, Frank J.; Nishimura, Tetsuo

    1993-01-01

    A large portion of the inner galactic plane has been mapped in the far-infrared (C II) line using a balloon-borne survey instrument. Complete coverage is reported from 25 degrees north to 80 degrees south of the galactic center and extending a few degrees on each side of the plane. Effective resolution is 14.1 acrmin (FWHM) and contour levels begin at 2 E -5 ergs/(s x sq. cm x ster). When compared with 100 micron dust emission observed by IRAS the (C II) appears well correlated with the dust emission except for a 10 degree region centered on the galactic center where emission from the gas is much weaker than that from the dust.

  1. Experience the natural sciences: Programs for teachers at the University of Hawaii at Hilo

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hapai, M.N.

    1994-12-31

    Since 1988, the University of Hawaii at Hilo Science and Education faculty have jointly created programs for pre- and in-service teachers, and to improve science teaching, to increase the number of science teachers, and to improve scientific literacy in the general population. The National Sciences major, approved in 1991, with both elementary and secondary teaching options, has gone from three degree seeking candidates in the fall of 1991 to fifty-nine in the spring of 1994. The major provides elementary teachers with a general science degree and teaching certification; and secondary teachers with a more intense general science degree, a specializedmore » minor, and teaching certification. Additionally, a new 18 credit Natural Sciences Certificate for in-service elementary teachers, designed to enhance their scientific background and classroom methodology, has already attracted over 250 teachers within the last year.« less

  2. CT measurement of range of motion of ankle and subtalar joints following two lateral column lengthening procedures.

    PubMed

    Beimers, Lijkele; Louwerens, Jan W K; Tuijthof, Gabrielle Josephine Maria; Jonges, Remmet; van Dijk, C N Niek; Blankevoort, Leendert

    2012-05-01

    Lateral column lengthening (LCL) has become an accepted procedure for the operative treatment of the flexible flatfoot deformity. Hindfoot arthrodesis via a calcaneocuboid distraction arthrodesis (CCDA) has been considered a less favourable surgical option than the anterior open wedge calcaneal distraction osteotomy (ACDO), as CCDA has been associated with reduced hindfoot joint motion postoperatively. The ankle and subtalar joint ranges of motion were measured in patients who underwent an ACDO or CCDA procedure for flatfoot deformity. CT scanning was performed with the foot in extreme positions in five ACDO and five CCDA patients. A bone segmentation and registration technique for the tibia, talus and calcaneus was applied to the CT images. Finite helical axis (FHA) rotations representing the range of motion of the joints were calculated for the motion between opposite extreme foot positions of the tibia and the calcaneus relative to the talus. The maximum mean FHA rotation of the ankle joint (for extreme dorsiflexion to extreme plantarflexion) after ACDO was 52.2 degrees ± 12.4 degrees and after CCDA 49.0 degrees ± 12.0 degrees. Subtalar joint maximum mean FHA rotation (for extreme eversion to extreme inversion) following ACDO was 22.8 degrees ± 8.6 degrees, and following CCDA 24.4 degrees ± 7.6 degrees. An accurate CT-based technique was used to assess the range of motion of the ankle and subtalar joints following two lateral column lengthening procedures for flatfoot deformity. Comparable results with a considerable amount of variance were found for the range of motion following the ACDO and CCDA procedures.

  3. Position controlled Knee Rehabilitation Orthotic Device for Patients after Total Knee Replacement Arthroplasty

    NASA Astrophysics Data System (ADS)

    Wannaphan, Patsiri; Chanthasopeephan, Teeranoot

    2016-11-01

    Knee rehabilitation after total knee replacement arthroplasty is essential for patients during their post-surgery recovery period. This study is about designing one degree of freedom knee rehabilitation equipment to assist patients for their post-surgery exercise. The equipment is designed to be used in sitting position with flexion/extension of knee in sagittal plane. The range of knee joint motion is starting from 0 to 90 degrees angle for knee rehabilitation motion. The feature includes adjustable link for different human proportions and the torque feedback control at knee joint during rehabilitation and the control of flexion/extension speed. The motion of the rehabilitation equipment was set to move at low speed (18 degrees/sec) for knee rehabilitation. The rehabilitation link without additional load took one second to move from vertical hanging up to 90° while the corresponding torque increased from 0 Nm to 2 Nm at 90°. When extra load is added, the link took 1.5 seconds to move to 90° The torque is then increased from 0 Nm to 4 Nm. After a period of time, the speed of the motion can be varied. User can adjust the motion to 40 degrees/sec during recovery activity of the knee and users can increase the level of exercise or motion up to 60 degrees/sec to strengthen the muscles during throughout their rehabilitation program depends on each patient. Torque control is included to prevent injury. Patients can use the equipment for home exercise to help reduce the number of hospital visit while the patients can receive an appropriate therapy for their knee recovery program.

  4. Motion deficit of the thumb in CMC joint arthritis.

    PubMed

    Gehrmann, Sebastian V; Tang, Jie; Li, Zong Ming; Goitz, Robert J; Windolf, Joachim; Kaufmann, Robert A

    2010-09-01

    Idiopathic osteoarthritis (OA) of the thumb carpometacarpal (CMC) joint is a common disabling disease that often causes pain and motion loss. The aims of this study were to characterize the multidimensional motion capability of the thumb CMC joint in a group with severe CMC OA and to compare it with a control group. We included 15 subjects with stage III/IV CMC OA according to the Eaton/Littler classification, and 15 control subjects. A motion analysis system using surface markers was employed to quantify the maximum boundary of the thumb circumduction envelope during repetitive thumb movements. We measured the area enclosed by the angular circumduction envelope and the ranges of motion (ROM) in multiple directions for the thumb CMC joint. Thumb osteoarthritis of the CMC joint stage III/IV resulted in a significantly smaller ROM in flexion/extension (45 degrees +/- 11 degrees for the CMC OA group, 59 degrees +/- 10 degrees for the controls), abduction-adduction (37 degrees +/- 6 degrees for the CMC OA group, 63 degrees +/- 13 degrees for the controls), and pronation-supination (49 degrees +/- 10 degrees for the CMC OA group, 62 degrees +/- 11 degrees for the controls) (p < .01). When analyzing the motion directions in flexion-extension and abduction-adduction separately, there was only a loss of extension and adduction (p < .01). Severe stages of thumb CMC OA cause an asymmetrical motion deficit with decreased ROM in extension and adduction, leading to decreased capability of counteropposition. Copyright 2010. Published by Elsevier Inc.

  5. Position of the prosthesis components in total ankle replacement and the effect on motion at the replaced joint.

    PubMed

    Cenni, Francesco; Leardini, Alberto; Cheli, Andrea; Catani, Fabio; Belvedere, Claudio; Romagnoli, Matteo; Giannini, Sandro

    2012-03-01

    In some cases of total ankle replacement, perfect alignment of the prosthetic components is not achieved. This study analyses the extent to which component positioning is critical for the final range of motion. Fourteen patients undergoing total ankle replacement were assessed preoperatively and postoperatively at seven and 13 months follow-up. X-ray pictures of the ankle were taken in static double leg stance, i.e. at neutral joint position, and in maximum plantarflexion and dorsiflexion. Measurements were obtained by a specially devised computer program based on anatomical reference points digitised on the radiograms. These allowed calculation of the position and orientation of the components in the sagittal and coronal planes, together with the joint range of motion. The mean range of motion was about 34 degrees at the first follow-up and maintained at the second. Tibial and talar components were more anterior than the mid-tibial shaft in 11 and nine patients, respectively. Mean inclination was about four degrees posterior for the tibial component and nearly one degree anterior for the talar component. A significantly larger range of motion was found in ankles both with the talar component located and inclined more anteriorly than the tibial. Correlation, though weak, was found between motion at the replaced ankle and possible residual subluxation and inclination of the components. However, a satisfactory range of motion was also achieved in those patients where recommended locations for the components could not be reached because of the size of the original joint deformity.

  6. One State's Response to the Collaborative Imperative: Voluntary Articulation Between the University of Wyoming and Wyoming's Community Colleges.

    ERIC Educational Resources Information Center

    Powell, Judith A.; Gillum, Forrest E.; Murdock, Maggi; Winter, Jerry; Muto, Jan

    1998-01-01

    Presents the statewide articulation agreements developed by Wyoming higher education institutions, resulting in a common name and numbering system for lower-division courses, as well as dual admission agreements; 2+2, 2+3, and 3+1 articulation agreements; joint delivery of degree programs; and shared data on student performance. These agreements…

  7. Implementing a Remote Laboratory Experience into a Joint Engineering Degree Program: Aerodynamic Levitation of a Beach Ball

    ERIC Educational Resources Information Center

    Jernigan, S. R.; Fahmy, Y.; Buckner, G. D.

    2009-01-01

    This paper details a successful and inexpensive implementation of a remote laboratory into a distance control systems course using readily available hardware and software. The physical experiment consists of a beach ball and a dc blower; the control objective is to make the height of the aerodynamically levitated beach ball track a reference…

  8. Integrated Data Modeling and Simulation on the Joint Polar Satellite System Program

    NASA Technical Reports Server (NTRS)

    Roberts, Christopher J.; Boyce, Leslye; Smith, Gary; Li, Angela; Barrett, Larry

    2012-01-01

    The Joint Polar Satellite System is a modern, large-scale, complex, multi-mission aerospace program, and presents a variety of design, testing and operational challenges due to: (1) System Scope: multi-mission coordination, role, responsibility and accountability challenges stemming from porous/ill-defined system and organizational boundaries (including foreign policy interactions) (2) Degree of Concurrency: design, implementation, integration, verification and operation occurring simultaneously, at multiple scales in the system hierarchy (3) Multi-Decadal Lifecycle: technical obsolesce, reliability and sustainment concerns, including those related to organizational and industrial base. Additionally, these systems tend to become embedded in the broader societal infrastructure, resulting in new system stakeholders with perhaps different preferences (4) Barriers to Effective Communications: process and cultural issues that emerge due to geographic dispersion and as one spans boundaries including gov./contractor, NASA/Other USG, and international relationships.

  9. Isokinetic and isometric strength-endurance after 6 hours of immersion and 6 degrees head-down tilt in men

    NASA Technical Reports Server (NTRS)

    Shaffer-Bailey, M.; Greenleaf, J. E.; Hutchinson, T. M.

    1996-01-01

    PURPOSE: To determine weight (water) loss levels for onset of muscular strength and endurance changes during deconditioning. METHODS: Seven men (27-40 yr) performed maximal shoulder-, knee-, and ankle-joint isometric (0 degree.s(-1) load) and isokinetic (60 degrees, 120 degrees, 180 degrees.s(-1) velocity) exercise tests during ambulatory control (AC), after 6 h of 6 degrees head-down tilt (HDT; dry-bulb temp. = 23.2 +/- SD 0.6 degrees C, relative humidity = 31.1+/- 11.1%) and after 6 h of 80 degrees foot-down head-out water immersion (WI; water temp. = 35.0 +/- SD 0.1 degree C) treatments. RESULTS: Weight (water) loss after HDT (1.10 +/- SE 0.14 kg, 1.4 +/- 0.2% body wt) and WI (1.54+/- 0.19 kg, 2.0 +/- 0.2% body wt) were not different, but urinary excretion with WI (1,354 +/- 142 ml.6 h(-1)) was 28% greater (p < 0.05) than that of 975 +/- 139 ml.6 h(-1) with HDT. Muscular endurance (total work; maximal flexion-extension of the non-dominant knee at 180 degrees.s(-1) for 30 s) was not different between AC and the WI or HDT treatments. Shoulder-, knee-, and ankle-joint strength was unchanged except for three knee-joint peak torques: AC torque (120 degrees.s(-1), 285 +/- 20 Nm) decreased to 268 +/- 21 Nm (delta = -6%, p < 0.05) with WI; and AC torques (180 degrees.s(-1), 260 +/- 19 Nm) decreased to 236 +/- 15 Nm (delta = -9%, p < 0.01) with HDT, and to 235 +/- 19 Nm (delta = -10%, p < 0.01) with WI. CONCLUSION: Thus, the total body hypohydration threshold level for shoulder- and ankle-joint strength and endurance decrements is more than 2% body weight (water) loss, while significant reduction in knee-joint muscular strength-endurance occurred only at moderate (120 degrees.s(-1) and lighter (180 degrees.s(-1)) loads with body weight loss of 1.4-2.0% following WI or HDT, respectively. These weight (water) losses and knee-joint strength decrements are somewhat less than the mean weight loss of 2.6% and knee-joint strength decrements of 6-20% of American astronauts after Skylab flights to 84 d.

  10. Effect of rotator cuff strengthening as an adjunct to standard care in subjects with adhesive capsulitis: A randomized controlled trial.

    PubMed

    Rawat, Pallavi; Eapen, Charu; Seema, Kulathuran Pillai

    Randomized controlled trial. To study the effect of adding rotator cuff (RC) muscles strengthening to joint mobilization and transcutaneous electrical nerve stimulation (TENS) in patients with adhesive capsulitis. A prospective, parallel-group, randomised clinical trial was conducted on 42 patients. One group received TENS and joint mobilization and in the other group RC muscles strengthening was added. Treatment was given for 12 sessions within 4 weeks. When compared between the groups statistically significant changes were seen in all the outcome measures in the group that received RC muscle strengthening exercises vs TENS and mobilization. VAS 12.76 ± 1.04 vs 4.05 ± 1.32; SPADI 34.66 ± 6.69 vs 54.29 ± 12.17; PFPS 3.06 ± 0.80 vs 4.70 ± 0.81; and ROM (elevation >125 vs >110 degrees and rotations >70 vs >48 degrees). Addition of a structured RC strengthening program to TENS and joint mobilization in the treatment of adhesive capsulitis resulted in improvement in pain, ROM and function. 1b. Copyright © 2016 Hanley & Belfus. Published by Elsevier Inc. All rights reserved.

  11. Credential Inflation and The Professional Doctorate in California Higher Education. Research & Occasional Paper Series: CSHE.1.04

    ERIC Educational Resources Information Center

    La Belle, Thomas J.

    2004-01-01

    The article argues that the time has come to change California's 1960 Master Plan for higher education by permitting the California State University (CSU) to award the doctorate in selected professional programs. The article also addresses the inadequacies of the joint doctorate as the means to remedy degree or credential creep; the CSU's focus on…

  12. Graduate engineering research participation in aeronautics

    NASA Technical Reports Server (NTRS)

    Roberts, A. S., Jr.

    1984-01-01

    Graduate student engineering research in aeronautics at Old Dominion University is surveyed. Student participation was facilitated through a NASA sponsored university program which enabled the students to complete degrees. Research summaries are provided and plans for the termination of the grant program are outlined. Project topics include: Failure modes for mechanically fastened joints in composite materials; The dynamic stability of an earth orbiting satellite deploying hinged appendages; The analysis of the Losipescu shear test for composite materials; and the effect of boundary layer structure on wing tip vortex formation and decay.

  13. Effectiveness of hydroplasty and therapeutic exercise for treatment of frozen shoulder.

    PubMed

    Callinan, Nancy; McPherson, Scott; Cleaveland, Susan; Voss, Debra Gardiner; Rainville, Darcel; Tokar, Nancy

    2003-01-01

    The purpose of this retrospective review was to evaluate the effectiveness of a hydraulic distention technique (hydroplasty) combined with a therapy program for treatment of idiopathic frozen shoulder. Over a two-year period, 60 patients with idiopathic frozen shoulder were identified as having undergone the hydroplasty procedure and therapy protocol at the authors' hand center. Distention of the glenohumeral joint was achieved by an injection of a 10-mL combination of bupivacaine (Marcaine), lidocaine (Xylocaine), and corticosteroid followed by injection of 30 mL of chilled sterile normal saline. Therapy was initiated immediately after the surgeon had completed the hydroplasty. The mean active range of motion improvement was as follows: flexion 28 degrees, abduction 42 degrees, internal rotation 22 degrees, and external rotation 26 degrees. There was no significant difference in outcomes between diabetics and nondiabetics or subjects with symptoms less than six months' duration compared with subjects with symptom duration greater than six months. At discharge, only two (3%) of the subjects reported persistent pain during sleep. The hydroplasty procedure combined with a therapy program is a successful treatment for idiopathic frozen shoulder.

  14. Custom electronic subsystems for the laboratory telerobotic manipulator

    NASA Technical Reports Server (NTRS)

    Glassell, R. L.; Butler, P. L.; Rowe, J. C.; Zimmermann, S. D.

    1990-01-01

    The National Aeronautics and Space Administration (NASA) Space Station Program presents new opportunities for the application of telerobotic and robotic systems. The Laboratory Telerobotic Manipulator (LTM) is a highly advanced 7 degrees-of-freedom (DOF) telerobotic/robotic manipulator. It was developed and built for the Automation Technology Branch at NASA's Langley Research Center (LaRC) for work in research and to demonstrate ground-based telerobotic manipulator system hardware and software systems for future NASA applications in the hazardous environment of space. The LTM manipulator uses an embedded wiring design with all electronics, motor power, and control and communication cables passing through the pitch-yaw differential joints. This design requires the number of cables passing through the pitch/yaw joint to be kept to a minimum. To eliminate the cables needed to carry each pitch-yaw joint's sensor data to the VME control computers, a custom-embedded electronics package for each manipulator joint was developed. The electronics package collects and sends the joint's sensor data to the VME control computers over a fiber optic cable. The electronics package consist of five individual subsystems: the VME Link Processor, the Joint Processor and the Joint Processor power supply in the joint module, the fiber optics communications system, and the electronics and motor power cabling.

  15. [Range of Hip Joint Motion and Weight of Lower Limb Function under 3D Dynamic Marker].

    PubMed

    Xia, Q; Zhang, M; Gao, D; Xia, W T

    2017-12-01

    To explore the range of reasonable weight coefficient of hip joint in lower limb function. When the hip joints of healthy volunteers under normal conditions or fixed at three different positions including functional, flexed and extension positions, the movements of lower limbs were recorded by LUKOtronic motion capture and analysis system. The degree of lower limb function loss was calculated using Fugl-Meyer lower limb function assessment form when the hip joints were fixed at the aforementioned positions. One-way analysis of variance and Tamhane's T2 method were used to proceed statistics analysis and calculate the range of reasonable weight coefficient of hip joint. There were significant differences between the degree of lower limb function loss when the hip joints fixed at flexed and extension positions and at functional position. While the differences between the degree of lower limb function loss when the hip joints fixed at flexed position and extension position had no statistical significance. In 95% confidence interval, the reasonable weight coefficient of hip joint in lower limb function was between 61.05% and 73.34%. Expect confirming the reasonable weight coefficient, the effects of functional and non-functional positions on the degree of lower limb function loss should also be considered for the assessment of hip joint function loss. Copyright© by the Editorial Department of Journal of Forensic Medicine

  16. A novel methodology to reproduce previously recorded six-degree of freedom kinematics on the same diarthrodial joint.

    PubMed

    Moore, Susan M; Thomas, Maribeth; Woo, Savio L-Y; Gabriel, Mary T; Kilger, Robert; Debski, Richard E

    2006-01-01

    The objective of this study was to develop a novel method to more accurately reproduce previously recorded 6-DOF kinematics of the tibia with respect to the femur using robotic technology. Furthermore, the effect of performing only a single or multiple registrations and the effect of robot joint configuration were investigated. A single registration consisted of registering the tibia and femur with respect to the robot at full extension and reproducing all kinematics while multiple registrations consisted of registering the bones at each flexion angle and reproducing only the kinematics of the corresponding flexion angle. Kinematics of the knee in response to an anterior (134 N) and combined internal/external (+/-10 N m) and varus/valgus (+/-5 N m) loads were collected at 0 degrees , 15 degrees , 30 degrees , 60 degrees , and 90 degrees of flexion. A six axes, serial-articulated robotic manipulator (PUMA Model 762) was calibrated and the working volume was reduced to improve the robot's accuracy. The effect of the robot joint configuration was determined by performing single and multiple registrations for three selected configurations. For each robot joint configuration, the accuracy in position of the reproduced kinematics improved after multiple registrations (0.7+/-0.3, 1.2+/-0.5, and 0.9+/-0.2 mm, respectively) when compared to only a single registration (1.3+/-0.9, 2.0+/-1.0, and 1.5+/-0.7 mm, respectively) (p<0.05). The accuracy in position of each robot joint configuration was unique as significant differences were detected between each of the configurations. These data demonstrate that the number of registrations and the robot joint configuration both affect the accuracy of the reproduced kinematics. Therefore, when using robotic technology to reproduce previously recorded kinematics, it may be necessary to perform these analyses for each individual robotic system and for each diarthrodial joint, as different joints will require the robot to be placed in different robot joint configurations.

  17. [Environmental Hazards Assessment Program annual report, June 1992--June 1993]. Proposal for a new program leading to the Master of Science degree in environmental studies to be offered jointly by the Medical University of South Carolina and the University of Charleston, South Carolina

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Not Available

    1993-12-01

    The Medical University of South Carolina (MUSC) and the University of Charleston, South Carolina (UCSC) propose to offer the degree of Master of Science in Environmental Studies. The proposed starting date is August 1994. The purpose of this interdisciplinary program is to offer nationally and internationally recognized graduate level training in the areas of environmental policy, science, and health risk assessment. Special emphasis will be placed on human health. Included in this proposal are a needs assessment for environmental science professionals along with employment projections and salary expectations. The Environmental Science program is described and its relationship to other programsmore » within MUSC and UCSC, as well as its relation to similar programs at other institutions are examined. Enrollment is discussed, admission requirements and standards outlined, and the curriculum is described. Academic and physical resources are examined and estimated costs are given.« less

  18. Strengthening cancer biology research, prevention, and control while reducing cancer disparities: student perceptions of a collaborative master's degree program in cancer biology, preventions, and control.

    PubMed

    Jillson, I A; Cousin, C E; Blancato, J K

    2013-09-01

    This article provides the findings of a survey of previous and current students in the UDC/GU-LCCC master's degree program. This master's degree program, Cancer Biology, Prevention, and Control is administered and taught jointly by faculty of a Minority Serving Institution, the University of the District of Columbia, and the Lombardi Comprehensive Cancer Center to incorporate the strengths of a community-based school with a research intensive medical center. The program was initiated in 2008 through agreements with both University administrations and funding from the National Cancer Institute. The master's degree program is 36 credits with a focus on coursework in biostatistics, epidemiology, tumor biology, cancer prevention, medical ethics, and cancer outreach program design. For two semesters during the second year, students work full-time with a faculty person on a laboratory or outreach project that is a requirement for graduation. Students are supported and encouraged to transition to a doctoral degree after they obtain the master's and many of them are currently in doctorate programs. Since the inception of the program, 45 students have initiated the course of study, 28 have completed the program, and 13 are currently enrolled in the program. The survey was designed to track the students in their current activities, as well as determine which courses, program enhancements, and research experiences were the least and most useful, and to discern students' perceptions of knowledge acquired on various aspects of Cancer Biology Prevention, and Control Master's Program. Thirty of the 35 individuals to whom email requests were sent responded to the survey, for a response rate of 85.7%. The results of this study will inform the strengthening of the Cancer Biology program by the Education Advisory Committee. They can also be used in the development of comparable collaborative master's degree programs designed to address the significant disparities in prevalence of cancer, low screening awareness, and access to and outcomes of cancer prevention and treatment services. This, in turn, will contribute to the elimination of the dearth of underrepresented minority scientists who address these disparities. By far, the students were satisfied with the program and believe that it has had significant impact on their ability to contribute to cancer prevention and control. They provided both general and specific recommendations to strengthen the program.

  19. The influence of foot position on stretching of the plantar fascia.

    PubMed

    Flanigan, Ryan M; Nawoczenski, Deborah A; Chen, Linlin; Wu, Hulin; DiGiovanni, Benedict F

    2007-07-01

    A recent study found nonweightbearing stretching exercises specific to the plantar fascia to be superior to the standard program of weightbearing Achilles tendon-stretching exercises in patients with chronic plantar fasciitis. The present study used a cadaver model to demonstrate the influence of foot and ankle position on stretching of the plantar fascia. Twelve fresh-frozen lower-leg specimens were tested in 15 different configurations representing various combinations of ankle and metatarsophalangeal (MTP) joint dorsiflexion, midtarsal transverse plane abduction and adduction, and forefoot varus and valgus. Measurements were recorded by a differential variable reluctance transducer (DVRT) implanted into the medial band of the plantar fascia, and primary measurement was a percent deformation of the plantar fascia (stretch) with respect to a reference position (90 degrees ankle dorsiflexion, 0 degrees midtarsal and forefoot orientation, and 0 degrees MTP dorsiflexion). Ankle and MTP joint dorsiflexion produced a significant increase (14.91%) in stretch compared to the position of either ankle dorsiflexion alone (9.31% increase, p < 0.001) or MTP dorsiflexion alone (7.33% increase, p < 0.01). There was no significant increase in stretch with positions of abduction or varus (2.49%, p = 0.27 and 0.55%, p = 0.79). This study provides a mechanical explanation for enhanced outcomes in recent clinical trials using plantar fascia tissue-specific stretching exercises and lends support to the use of ankle and MTP joint dorsiflexion when employing stretching protocols for nonoperative treatment in patients with chronic proximal plantar fasciitis.

  20. Long-term neuromuscular training and ankle joint position sense.

    PubMed

    Kynsburg, A; Pánics, G; Halasi, T

    2010-06-01

    Preventive effect of proprioceptive training is proven by decreasing injury incidence, but its proprioceptive mechanism is not. Major hypothesis: the training has a positive long-term effect on ankle joint position sense in athletes of a high-risk sport (handball). Ten elite-level female handball-players represented the intervention group (training-group), 10 healthy athletes of other sports formed the control-group. Proprioceptive training was incorporated into the regular training regimen of the training-group. Ankle joint position sense function was measured with the "slope-box" test, first described by Robbins et al. Testing was performed one day before the intervention and 20 months later. Mean absolute estimate errors were processed for statistical analysis. Proprioceptive sensory function improved regarding all four directions with a high significance (p<0.0001; avg. mean estimate error improvement: 1.77 degrees). This was also highly significant (p< or =0.0002) in each single directions, with avg. mean estimate error improvement between 1.59 degrees (posterior) and 2.03 degrees (anterior). Mean absolute estimate errors at follow-up (2.24 degrees +/-0.88 degrees) were significantly lower than in uninjured controls (3.29 degrees +/-1.15 degrees) (p<0.0001). Long-term neuromuscular training has improved ankle joint position sense function in the investigated athletes. This joint position sense improvement can be one of the explanations for injury rate reduction effect of neuromuscular training.

  1. [The retrocapital osteotomy ("chevron") for correction of splayfoot with hallux valgus].

    PubMed

    Gabel, Michael

    2008-12-01

    Surgical treatment of hallux valgus deformity with a distal osteotomy of the first metatarsal to address an increased intermetatarsal angle (IMA) I-II. This procedure is combined with a soft-tissue procedure at the first metatarsophalangeal joint: realignment of the first ray, lateral displacement of the first metatarsal head above the sesamoids, rebalancing of the soft tissues at the metatarsophalangeal joint. Pain and soft-tissue inflammation at the bunion, impaired function of the metatarsophalangeal joint, and lateral deviation of the hallux. IMA I-II 10 degrees. Symptomatic osteoarthritis of the first metatarsophalangeal joint, assessed clinically or radiographically. Acute inflammation of the forefoot, osteoporosis of the first metatarsal. Vascular disturbance. Cosmetic indication only. Relative: hypermobility of the first ray, valgus malalignment of the hindfoot, previous retrocapital osteotomy. Lateral soft-tissue release. Resection of the medial pseudoexostosis. V-shaped osteotomy of the distal metatarsal I. Exostosectomy. Lateral displacement of the first metatarsal head. Screw fixation. Realignment of the metatarsophalangeal joint by tightening of the medial soft tissues. Postoperative shoe with full weight bearing. Active exercises of the foot and hallux. Physiotherapy. Prophylaxis of deep vein thrombosis depending on the degree of mobility. Radiographic control after 6 weeks. Bandage or orthosis to maintain toe alignment. IMA I-II was reduced from 13.6 degrees preoperatively to 6.6 degrees postoperatively. HVA decreased from 29.8 degrees to 8.2 degrees postoperatively.

  2. Hip and knee joints are more stabilized than driven during the stance phase of gait: an analysis of the 3D angle between joint moment and joint angular velocity.

    PubMed

    Dumas, R; Cheze, L

    2008-08-01

    Joint power is commonly used in orthopaedics, ergonomics or sports analysis but its clinical interpretation remains controversial. Some basic principles on muscle actions and energy transfer have been proposed in 2D. The decomposition of power on 3 axes, although questionable, allows the same analysis in 3D. However, these basic principles have been widely criticized, mainly because bi-articular muscles must be considered. This requires a more complex computation in order to determine how the individual muscle force contributes to drive the joint. Conversely, with simple 3D inverse dynamics, the analysis of both joint moment and angular velocity directions is essential to clarify when the joint moment can contribute or not to drive the joint. The present study evaluates the 3D angle between the joint moment and the joint angular velocity and investigates when the hip, knee and ankle joints are predominantly driven (angle close to 0 degrees and 180 degrees ) or stabilized (angle close to 90 degrees ) during gait. The 3D angle curves show that the three joints are never fully but only partially driven and that the hip and knee joints are mainly stabilized during the stance phase. The notion of stabilization should be further investigated, especially for subjects with motion disorders or prostheses.

  3. Airsea Battle: Can the Air Force and the Navy Get Along

    DTIC Science & Technology

    2011-06-01

    staffs at headquarters, Special Operations Command – Korea (SOCKOR) and at Joint Special Operations Task Force – Philippines (JSOTF- P ). His most...Detachment (JSOAD) deputy commander for JSOTF- P . Major Kobs holds a Bachelors of Arts Degree in Political Science from the University of New...1894), p . 26. concept, Glenn H. Curtiss agreed to instruct naval officers for the creation of a naval aviation program. Lieutenant T.G

  4. Tibiofemoral joint contact area and pressure after single- and double-bundle anterior cruciate ligament reconstruction.

    PubMed

    Morimoto, Yusuke; Ferretti, Mario; Ekdahl, Max; Smolinski, Patrick; Fu, Freddie H

    2009-01-01

    The purpose of this study was to compare the tibiofemoral contact area and pressure after single-bundle (SB) and double-bundle (DB) anterior cruciate ligament (ACL) reconstruction by use of 2 femoral and 2 tibial tunnels in intact cadaveric knees. Tibiofemoral contact area and mean and maximum pressures were measured by pressure-sensitive film (Fujifilm, Valhalla, NY) inserted between the tibia and femur. The knee was subjected to a 1,000-N axial load by use of a uniaxial testing machine at 0 degrees , 15 degrees , 30 degrees , and 45 degrees of flexion. Three conditions were evaluated: (1) intact ACL, (2) SB ACL reconstruction (n = 10 knees), and (3) DB ACL reconstruction (n = 9 knees). When compared with the intact knee, DB ACL reconstruction showed no significant difference in tibiofemoral contact area and mean and maximum pressures. SB ACL reconstruction had a significantly smaller contact area on the lateral and medial tibiofemoral joints at 30 degrees and 15 degrees of flexion. SB ACL reconstruction also had significantly higher mean pressures at 15 degrees of flexion on the medial tibiofemoral joint and at 0 degrees and 15 degrees of flexion on the lateral tibiofemoral joint, as well as significantly higher maximum pressures at 15 degrees of flexion on the lateral tibiofemoral joint. SB ACL reconstruction resulted in a significantly smaller tibiofemoral contact area and higher pressures. DB ACL more closely restores the normal contact area and pressure mainly at low flexion angles. Our findings suggest that the changes in the contact area and pressures after SB ACL reconstruction may be one of the causes of osteoarthritis on long-term follow-up. DB ACL reconstruction may reduce the incidence of osteoarthritis by closely restoring contact area and pressure.

  5. Mechanical flexible joint design document

    NASA Technical Reports Server (NTRS)

    Daily, Vic

    1993-01-01

    The purpose of this report is to document the status of the Mechanical Flexible Joint (MFJ) Design Subtask with the intent of halting work on the design. Recommendations for future work is included in the case that the task is to be resumed. The MFJ is designed to eliminate two failure points from the current flex joint configuration, the inner 'tripod configuration' and the outer containment jacket. The MFJ will also be designed to flex 13.5 degrees and have three degrees of freedom. By having three degrees of freedom, the MFJ will allow the Low Pressure Fuel Duct to twist and remove the necessity to angulate the full 11 degrees currently required. The current flex joints are very labor intensive and very costly and a simple alternative is being sought. The MFJ is designed with a greater angular displacement, with three degrees of freedom, to reside in the same overall envelope, to meet weight constraints of the current bellows, to be compatible with cryogenic fuel and oxidizers, and also to be man-rated.

  6. Descending and Local Network Interactions Control Adaptive Locomotion

    DTIC Science & Technology

    2014-12-04

    segment). The thorax-coxa ( ThC ) joint swings dramatically in the front leg, but hardly moves at all in the middle and hind legs. These observations...changes in the three degree of freedom thorax-coxa joint ( ThC ) (Fig. 1). Indeed we were able to describe which of the three degrees of freedom were...these actions are distinct. The T1 leg relies heavily on actions of the second and third degree of freedom of the ThC joint ( ThC 2 and ThC3) whereas in

  7. Dizziness and unsteadiness following whiplash injury: characteristic features and relationship with cervical joint position error.

    PubMed

    Treleaven, Julia; Jull, Gwendolen; Sterling, Michele

    2003-01-01

    Dizziness and/or unsteadiness are common symptoms of chronic whiplash-associated disorders. This study aimed to report the characteristics of these symptoms and determine whether there was any relationship to cervical joint position error. Joint position error, the accuracy to return to the natural head posture following extension and rotation, was measured in 102 subjects with persistent whiplash-associated disorder and 44 control subjects. Whiplash subjects completed a neck pain index and answered questions about the characteristics of dizziness. The results indicated that subjects with whiplash-associated disorders had significantly greater joint position errors than control subjects. Within the whiplash group, those with dizziness had greater joint position errors than those without dizziness following rotation (rotation (R) 4.5 degrees (0.3) vs 2.9 degrees (0.4); rotation (L) 3.9 degrees (0.3) vs 2.8 degrees (0.4) respectively) and a higher neck pain index (55.3% (1.4) vs 43.1% (1.8)). Characteristics of the dizziness were consistent for those reported for a cervical cause but no characteristics could predict the magnitude of joint position error. Cervical mechanoreceptor dysfunction is a likely cause of dizziness in whiplash-associated disorder.

  8. Altered neuromuscular control and ankle joint kinematics during walking in subjects with functional instability of the ankle joint.

    PubMed

    Delahunt, Eamonn; Monaghan, Kenneth; Caulfield, Brian

    2006-12-01

    The ankle joint requires very precise neuromuscular control during the transition from terminal swing to the early stance phase of the gait cycle. Altered ankle joint arthrokinematics and muscular activity have been cited as potential factors that may lead to an inversion sprain during the aforementioned time periods. However, to date, no study has investigated patterns of muscle activity and 3D joint kinematics simultaneously in a group of subjects with functional instability compared with a noninjured control group during these phases of the gait cycle. To compare the patterns of lower limb 3D joint kinematics and electromyographic activity during treadmill walking in a group of subjects with functional instability with those observed in a control group. Controlled laboratory study. Three-dimensional angular velocities and displacements of the hip, knee, and ankle joints, as well as surface electromyography of the rectus femoris, peroneus longus, tibialis anterior, and soleus muscles, were recorded simultaneously while subjects walked on a treadmill at a velocity of 4 km/h. Before heel strike, subjects with functional instability exhibited a decrease in vertical foot-floor clearance (12.62 vs 22.84 mm; P < .05), as well as exhibiting a more inverted position of the ankle joint before, at, and immediately after heel strike (1.69 degrees , 2.10 degrees , and -0.09 degrees vs -1.43 degrees , -1.43 degrees , and -2.78 degrees , respectively [minus value = eversion]; P < .05) compared with controls. Subjects with functional instability were also observed to have an increase in peroneus longus integral electromyography during the post-heel strike time period (107.91%.millisecond vs 64.53%.millisecond; P < .01). The altered kinematics observed in this study could explain the reason subjects with functional instability experience repeated episodes of ankle inversion injury in situations with only slight or no external provocation. It is hypothesized that the observed increase in peroneus longus activity may be the result of a change in preprogrammed feed-forward motor control.

  9. Cranial cruciate stability in the rottweiler and racing greyhound: an in vitro study.

    PubMed

    Wingfield, C; Amis, A A; Stead, A C; Law, H T

    2000-05-01

    An in vitro biomechanical study of cadaver stifles from rottweilers and racing greyhounds was undertaken to evaluate the contribution of the cranial cruciate ligament to stifle joint stability. This was performed at differing stifle joint angles, first with the joint capsules and ligaments intact and then with all structures removed except for the cranial cruciate ligament. Craniocaudal laxity increased in both breeds as stifle flexion increased. The rottweiler stifle showed greater craniocaudal joint laxity than the racing greyhound at all joint angles between 150 degrees and 110 degrees, but the actual increases in joint laxity between these joint angles were similar for both breeds. Tibial rotation during craniocaudal loading of the stifle increased craniocaudal laxity in both breeds during joint flexion. The relative contribution of the cranial cruciate ligament to cranial stability of the stifle joint increased as the joint flexed and was similar in both breeds.

  10. Equatorial Wave Line, Pacific Ocean

    NASA Image and Video Library

    1993-01-19

    STS054-95-042 (13-19 Jan 1993) --- The Equatorial Pacific Ocean is represented in this 70mm view. The international oceanographic research community is presently conducting a program called Joint Global Ocean Flux Study (JGOFS) to study the global ocean carbon budget. A considerable amount of effort within this program is presently being focused on the Equatorial Pacific Ocean because of the high annual average biological productivity. The high productivity is the result of nearly constant easterly winds causing cool, nutrient-rich water to well up at the equator. In this view of the sun glint pattern was photographed at about 2 degrees north latitude, 103 degrees west longitude, as the Space Shuttle passed over the Equatorial Pacific. The long narrow line is the equatorial front, which defines the boundary between warm surface equatorial water and cool, recently upwelled water. Such features are of interest to the JGOFS researchers and it is anticipated that photographs such as this will benefit the JGOFS program.

  11. Ankle rehabilitation device with two degrees of freedom and compliant joint

    NASA Astrophysics Data System (ADS)

    Racu (Cazacu, C.-M.; Doroftei, I.

    2015-11-01

    We propose a rehabilitation device that we intend to be low cost and easy to manufacture. The system will ensure functionality but also have a small dimensions and low mass, considering the physiological dimensions of the foot and lower leg. To avoid injure of the ankle joint, this device is equipped with a compliant joint between the motor and mechanical transmission. The torque of this joint is intended to be adjustable, according to the degree of ankle joint damage. To choose the material and the dimensions of this compliant joint, in this paper we perform the first stress simulation. The minimum torque is calculated, while the maximum torque is given by the preliminary chosen actuator.

  12. Civil-Military Relations: From Vietnam to Operation Iraqi Freedom

    DTIC Science & Technology

    2009-05-01

    Robert F. Baumann, Ph.D. Graduate Degree Programs ii Abstract CIVIL-MILITARY RELATIONS: FROM VIETNAM TO OPERATION IRAQI FREEDOM by MAJOR Brandon...Oxford: Oxford University Press, 2007), 285-324. 43 Ibid. 44 Robert M. Gates, 2008 National Defense Strategy (Washington D.C.: Department Of Defense...Dereliction of Duty: Lyndon Johnson, Robert McNamara, the Joint Chiefs of Staff, and the Lies That Led to Vietnam (New York: HarperCollins, 1997), 13. 60

  13. Historical overview of Intercosmos program

    NASA Astrophysics Data System (ADS)

    Rimsha, M. A.

    1986-02-01

    Fifteen years ago, on October 14, 1969, the artificial Earth satellite Intercosmos-1 was launched initiating joint satellite research conducted by scientists and specialists from friendly countries. A little more than 2 years were spent on the preparations for this launch. The first envoys of the Intercosmos program were comparatively small space vehicles. They consisted of three basic sections: a cylindrical middle section and two hemispheres. The first craft of the Intercosmos series achieved orientation on the Sun with an accuracy of a few angular degrees. The launch of the Intercosmos-1 satellite initiated one of the basic directions of research conducted in accordance with the program of space cooperation by the socialist countries--research on the solar-terrestrial ties using satellites.

  14. Prevalence and inheritance of and selection for hip dysplasia in seven breeds of dogs in Sweden and benefit: cost analysis of a screening and control program.

    PubMed

    Swenson, L; Audell, L; Hedhammar, A

    1997-01-15

    To determine the prevalence and changes over time in the prevalence of hip dysplasia; to ascertain whether prevalence or severity of hip dysplasia was associated with sex of the dogs, age at which coxofemoral joint status was evaluated, or ancestral background; to determine the effects of selective breeding; and to conduct an economic evaluation of the hip dysplasia program operated by the Swedish Kennel Club. Analysis of radiographic evaluations of coxofemoral joint conformity. 83,229 dogs from 7 breeds registered by the Swedish Kennel Club. All radiographs were scrutinized by a single radiologist (LA), and coxofemoral joint conformation was classified as normal or dysplastic, with the degree of dysplasia classified as 1,2,3, or 4. Decreasing prevalence of hip dysplasia corresponding to selection of breeding stock and high heritabilities was found. Sex differences were documented in 3 of the breeds. This was interpreted as breed differences in the distribution of genes related to hip dysplasia. Economic analyses showed that costs of screening and registration of coxofemoral joints was less than the value of dogs estimated to have been saved from moderate, severe, or very severe hip dysplasia in 6 of the breeds. Documented effects of age suggest that all dogs should be screened at the same age, rather than screening a few dogs at an older, more revealing age. In screening and control programs based on an open registry with access to family records, decreasing prevalence of hip dysplasia can be expected, and related to selection of breeding stock.

  15. Effects of footwear on lead limb knee and ankle joint kinematics in a fast bowler with a history of posterior ankle joint impingement-a case report.

    PubMed

    Bishop, Chris; Bartold, Simon; Thewlis, Dominic

    2013-11-01

    This case study reports the kinematic effect of 2 different cricket shoes on a fast bowler who reports a history of posterior ankle joint impingement. The participant bowled 6 trials in 2 pairs of cricket shoes. The 3-dimensional kinematics of the joints of the front leg was quantified during stance phase of the delivery stride. Wearing the high-cut shoe resulted in the ankle being 7.7-degree angle more plantarflexed at initial contact compared with the low-cut shoe. Again, when wearing the high-cut shoe compared with the low-cut shoe, the ankle joint was 15.5-degree angle more adducted and the knee was 4.1-degree angle less externally rotated at initial contact. This case study identifies the bowler's preferred shoe (high-cut shoe) as a potential contributing factor to the symptoms he was experiencing.

  16. Decentralized control of large flexible structures by joint decoupling

    NASA Technical Reports Server (NTRS)

    Su, Tzu-Jeng; Juang, Jer-Nan

    1994-01-01

    This paper presents a novel method to design decentralized controllers for large complex flexible structures by using the idea of joint decoupling. Decoupling of joint degrees of freedom from the interior degrees of freedom is achieved by setting the joint actuator commands to cancel the internal forces exerting on the joint degrees of freedom. By doing so, the interactions between substructures are eliminated. The global structure control design problem is then decomposed into several substructure control design problems. Control commands for interior actuators are set to be localized state feedback using decentralized observers for state estimation. The proposed decentralized controllers can operate successfully at the individual substructure level as well as at the global structure level. Not only control design but also control implementation is decentralized. A two-component mass-spring-damper system is used as an example to demonstrate the proposed method.

  17. Can symptomatic acromioclavicular joints be differentiated from asymptomatic acromioclavicular joints on 3-T MR imaging?

    PubMed

    Choo, Hye Jung; Lee, Sun Joo; Kim, Jung Han; Cha, Seong Sook; Park, Young Mi; Park, Ji Sung; Lee, Jun Woo; Oh, Minkyung

    2013-04-01

    To evaluate retrospectively whether symptomatic acromioclavicular joints can be differentiated from asymptomatic acromioclavicular joints on 3-T MR imaging. This study included 146 patients who underwent physical examination of acromioclavicular joints and 3-T MR imaging of the shoulder. Among them, 67 patients showing positive results on physical examination were assigned to the symptomatic group, whereas 79 showing negative results were assigned to the asymptomatic group. The following MR findings were compared between the symptomatic and asymptomatic groups: presence of osteophytes, articular surface irregularity, subchondral cysts, acromioclavicular joint fluid, subacromial fluid, subacromial bony spurs, joint capsular distension, bone edema, intraarticular enhancement, periarticular enhancement, superior and inferior joint capsular distension degree, and joint capsular thickness. The patients were subsequently divided into groups based on age (younger, older) and the method of MR arthrography (direct MR arthrography, indirect MR arthrography), and all the MR findings in each subgroup were reanalyzed. The meaningful cutoff value of each significant continuous variable was calculated using receiver operating characteristic analysis. The degree of superior capsular distension was the only significant MR finding of symptomatic acromioclavicular joints and its meaningful cutoff value was 2.1mm. After subgroup analyses, this variable was significant in the older age group and indirect MR arthrography group. On 3-T MR imaging, the degree of superior joint capsular distension might be a predictable MR finding in the diagnosis of symptomatic acromioclavicular joints. Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

  18. Simultaneous, Joint Inversion of Seismic Body Wave Travel Times and Satellite Gravity Data for Three-Dimensional Tomographic Imaging of Western Colombia

    NASA Astrophysics Data System (ADS)

    Dionicio, V.; Rowe, C. A.; Maceira, M.; Zhang, H.; Londoño, J.

    2009-12-01

    We report on the three-dimensional seismic structure of western Colombia determined through the use of a new, simultaneous, joint inversion tomography algorithm. Using data recorded by the national Seismological Network of Colombia (RSNC), we have selected 3,609 earthquakes recorded at 33 sensors distributed throughout the country, with additional data from stations in neighboring countries. 20,338 P-wave arrivals and 17,041 S-wave arrivals are used to invert for structure within a region extending approximately 72.5 to 77.5 degrees West and 2 to 7.5 degrees North. Our algorithm is a modification of the Maceira and Ammon joint inversion code, in combination with the Zhang and Thurber TomoDD (double-difference tomography) program, with a fast LSQR solver operating on the gridded values jointly. The inversion uses gravity anomalies obtained during the GRACE2 satellite mission, and solves using these values with the seismic travel-times through application of an empirical relationship first proposed by Harkrider, mapping densities to Vp and Vs within earth materials. In previous work, Maceira and Ammon demonstrated that incorporation of gravity data predicts shear wave velocities more accurately than the inversion of surface waves alone, particularly in regions where the crust exhibits abrupt and significant lateral variations in lithology, such as the Tarim Basin. The significant complexity of crustal structure in Colombia, due to its active tectonic environment, makes it a good candidate for the application with gravity and body waves. We present the results of this joint inversion and compare it to results obtained using travel times alone

  19. Chondromalacia as pathological finding in arthroscopy of the temporomandibular joint: A retrospective study.

    PubMed

    Martin-Granizo, Rafael; Correa-Muñoz, Diana Carolina

    2018-01-01

    The objective of this study was to describe the arthroscopic findings of chondromalacia and its relation with the internal derangement of the temporomandibular joint (TMJ). A total of 161 patients (299 TMJs) who underwent arthroscopy were included in the study. The TMJs were evaluated objectively under arthroscopic vision, and 4 groups of patients were established according to the degree of involvement, degree I, II, III and IV. Statistical analyses were conducted using logistic regression models (P < 0.05). It was observed that 95 patients (59%) had no sign of chondromalacia and 66 (41%) in 88 joints exhibited some degree of chondromalacia (44 patients unilaterally and 22 bilaterally). Of the 88 joints with chondromalacia, 14 (15.9%) had chondromalacia degree I, 12 (13.6%) chondromalacia degree II, 20 (22.7%) chondromalacia degree III and 42 (47.7%) chondromalacia degree IV. The chondromalacia was more significantly found in patients with ADDwR and discal perforation (P < 0.05), even as a common finding in patients without any internal deragement. Chondromalacia degree IV was a significant finding in cases of ADDwoR (P = 0.000619). Chondromalacia of the TMJ is a common finding in patients with internal derangement even at the early stages. Copyright © 2016 European Association for Cranio-Maxillo-Facial Surgery. Published by Elsevier Ltd. All rights reserved.

  20. New master program in management in biophotonics and biotechnologies

    NASA Astrophysics Data System (ADS)

    Meglinski, I. V.; Tuchin, V. V.

    2006-08-01

    We develop new graduate educational highly interdisciplinary program that will be useful for addressing problems in worldwide biotechnologies and related biomedical industries. This Master program called Management in Biophotonics and Biotechnologies provides students with the necessary training, education and problem-solving skills to produce managers who are better equipped to handle the challenges of modern business in modern biotechnologies. Administered jointly by Cranfield University (UK) and Saratov State University, Russia) graduates possess a blend of engineering, biotechnologies, business and interpersonal skills necessary for success in industry. The Master courses combine a regular year program in biophotonics & biotechnologies disciplines with the core requirements of a Master degree. A major advantage of the program is that it will provide skills not currently available to graduates in any other program, and it will give the graduates an extra competitive edge for getting a job then.

  1. The potential role of brain asymmetry in the development of adolescent idiopathic scoliosis: a hypothesis.

    PubMed

    Niesluchowski, W; Dabrowska, A; Kedzior, K; Zagrajek, T

    1999-10-01

    The size asymmetry of cerebral hemispheres may predispose to head tilt and asymmetric blocking of the zygapophysial joints, potentially leading to the development of compensatory curvatures in the lower segments of the spine. To analyze the effects of spinal manipulation, maintained by an exercise program, on the progression of idiopathic adolescent scoliosis in 2 children aged 6 and 10. The scoliosis found was 16 and 60 degrees. For diagnosis and monitoring of therapy, we recorded qualitative parameters of shoulder asymmetry, axillary line asymmetry, and scapular angle position. Manual treatment consisted of the examinations of all sliding motion in zygapophysial joints and both sacroiliac joints and removing the limitations of the sliding motions according to the method of Karel Lewit. The treatment procedure consisted of 3 or 4 manipulations within 17 months and an exercise program. The manipulation effects were maintained by the exercise program. The exercises were done in 2 or 3 sessions weekly for a year. In both patients we observed that scoliosis decompensation was successfully stopped and the effects of the correction persisted for 10 years. Brain and head asymmetry may be only a transient state, predisposing to asymmetric blocking at the atlanto-occipital level. Removal of blocking may prevent curve progression in children who had adolescent idiopathic scoliosis. The manipulative therapy may also have a promising effect on retarding curve progression when used in skeletally immature patient.

  2. Strengthening Cancer Biology Research, Prevention, and Control while Reducing Cancer Disparities: Student Perceptions of a Collaborative Master’s Degree Program in Cancer Biology, Preventions, and Control

    PubMed Central

    Cousin, C. E.; Blancato, J. K.

    2018-01-01

    This article provides the findings of a survey of previous and current students in the UDC/GU-LCCC master’s degree program. This master’s degree program, Cancer Biology, Prevention, and Control is administered and taught jointly by faculty of a Minority Serving Institution, the University of the District of Columbia, and the Lombardi Comprehensive Cancer Center to incorporate the strengths of a community-based school with a research intensive medical center. The program was initiated in 2008 through agreements with both University administrations and funding from the National Cancer Institute. The master’s degree program is 36 credits with a focus on coursework in biostatistics, epidemiology, tumor biology, cancer prevention, medical ethics, and cancer outreach program design. For two semesters during the second year, students work full-time with a faculty person on a laboratory or outreach project that is a requirement for graduation. Students are supported and encouraged to transition to a doctoral degree after they obtain the master’s and many of them are currently in doctorate programs. Since the inception of the program, 45 students have initiated the course of study, 28 have completed the program, and 13 are currently enrolled in the program. The survey was designed to track the students in their current activities, as well as determine which courses, program enhancements, and research experiences were the least and most useful, and to discern students’ perceptions of knowledge acquired on various aspects of Cancer Biology Prevention, and Control Master’s Program. Thirty of the 35 individuals to whom email requests were sent responded to the survey, for a response rate of 85.7 %. The results of this study will inform the strengthening of the Cancer Biology program by the Education Advisory Committee. They can also be used in the development of comparable collaborative master’s degree programs designed to address the significant disparities in prevalence of cancer, low screening awareness, and access to and outcomes of cancer prevention and treatment services. This, in turn, will contribute to the elimination of the dearth of underrepresented minority scientists who address these disparities. By far, the students were satisfied with the program and believe that it has had significant impact on their ability to contribute to cancer prevention and control. They provided both general and specific recommendations to strengthen the program. PMID:23784366

  3. Effect of isokinetic training on strength, functionality and proprioception in athletes with functional ankle instability.

    PubMed

    Sekir, Ufuk; Yildiz, Yavuz; Hazneci, Bulent; Ors, Fatih; Aydin, Taner

    2007-05-01

    The purpose of this study was to investigate the effects of isokinetic exercise on strength, joint position sense and functionality in recreational athletes with functional ankle instability (FAI). Strength, proprioception and balance of 24 recreational athletes with unilateral FAI were evaluated by using isokinetic muscle strength measurement, ankle joint position sense and one leg standing test. The functional ability was evaluated using five different tests. These were; single limb hopping course (SLHC), one legged and triple legged hop for distance (OLHD-TLHD), and six and cross six meter hop for time (SMHT-CSMHT). Isokinetic peak torque of the ankle invertor and evertor muscles were assessed eccentrically and concentrically at test speeds of 120 degrees /s. Isokinetic exercise protocol was carried out at an angular velocity of 120 degrees /s. The exercise session was repeated three times a week and lasted after 6 weeks. At baseline, concentric invertor strength was found to be significantly lower in the functionally unstable ankles compared to the opposite healthy ankles (p < 0.001). This difference was not present after executing the 6 weeks exercise sessions (p > 0.05). Ankle joint position sense in the injured ankles declined significantly from 2.35 +/- 1.16 to 1.33 +/- 0.62 degrees for 10 degrees of inversion angle (p < 0.001) and from 3.10 +/- 2.16 to 2.19 +/- 0.98 degrees for 20 degrees of inversion angle (p < 0.05) following the isokinetic exercise. One leg standing test score decreased significantly from 15.17 +/- 8.50 to 11.79 +/- 7.81 in the injured ankles (p < 0.001). Following the isokinetic exercise protocol, all of the worsened functional test scores in the injured ankles as compared to the opposite healthy ankles displayed a significant improvement (p < 0.01 for OLHD and CSMHT, p < 0.001 for SLHC, TLHD, and SMHT). These results substantiate the deficits of strength, proprioception, balance and functionality in recreational athletes with FAI. The isokinetic exercise program used in this study had a positive effect on these parameters.

  4. Femur rotation and patellofemoral joint kinematics: a weight-bearing magnetic resonance imaging analysis.

    PubMed

    Souza, Richard B; Draper, Christie E; Fredericson, Michael; Powers, Christopher M

    2010-05-01

    Controlled laboratory study using a cross-sectional design. To compare patellofemoral joint kinematics, femoral rotation, and patella rotation between females with patellofemoral pain (PFP) and pain-free controls using weight-bearing kinematic magnetic resonance imaging. Recently, it has been recognized that patellofemoral malalignment may be the result of femoral motion as opposed to patella motion. Fifteen females with PFP and 15 pain-free females between the ages of 18 and 45 years participated in this study. Kinematic imaging of the patellofemoral joint was performed using a vertically open magnetic resonance imaging system. Axial-oblique images were obtained using a fast gradient-echo pulse sequence. Images were acquired at a rate of 1 image per second while subjects performed a single-limb squat. Measures of femur and patella rotation (relative to the image field of view), lateral patella tilt, and lateral patella displacement were made from images obtained at 45 degrees , 30 degrees , 15 degrees , and 0 degrees of knee flexion. Group differences were assessed using a mixed-model analysis of variance with repeated measures. When compared to the control group, females with PFP demonstrated significantly greater lateral patella displacement at all angles evaluated and significantly greater lateral patella tilt at 30 degrees , 15 degrees , and 0 degrees of knee flexion. Similarly, greater medial femoral rotation was observed in the PFP group at 45 degrees , 15 degrees , and 0 degrees of knee flexion when compared to the control group. No group differences in patella rotation were found. Altered patellofemoral joint kinematics in females with PFP appears to be related to excessive medial femoral rotation, as opposed to lateral patella rotation. Our results suggest that the control of femur rotation may be important in restoring normal patellofemoral joint kinematics. J Orthop Sports Phys Ther 2010;40(5):277-285, Epub 12 March 2010. doi:10.2519/jospt.2010.3215.

  5. Three Degree of Freedom Parallel Mechanical Linkage

    NASA Technical Reports Server (NTRS)

    Adelstein, Bernard D. (Inventor)

    1998-01-01

    A three degree of freedom parallel mechanism or linkage that couples three degree of freedom translational displacements at an endpoint, such as a handle, a hand grip, or a robot tool, to link rotations about three axes that are fixed with respect to a common base or ground link. The mechanism includes a three degree of freedom spherical linkage formed of two closed loops, and a planar linkage connected to the endpoint. The closed loops are rotatably interconnected, and made of eight rigid links connected by a plurality of single degree of freedom revolute joints. Three of these revolute joints are base joints and are connected to a common ground. such that the axis lines passing through the revolute joints intersect at a common fixed center point K forming the center of a spherical work volume in which the endpoint is capable of moving. 'Me three degrees of freedom correspond to the spatial displacement of the endpoint, for instance. The mechanism provides a new overall spatial kinematic linkage composed of a minimal number of rigid links and rotary joints. The mechanism has improved mechanical stiffness, and conveys mechanical power bidirectionally between the human operator and the electromechanical actuators. It does not require gears, belts. cable, screw or other types of transmission elements, and is useful in applications requiring full backdrivability. Thus, this invention can serve as the mechanical linkage for actively powered devices such as compliant robotic manipulators and force-reflecting hand controllers, and passive devices such as manual input devices for computers and other systems.

  6. Fuzzy logic control of telerobot manipulators

    NASA Technical Reports Server (NTRS)

    Franke, Ernest A.; Nedungadi, Ashok

    1992-01-01

    Telerobot systems for advanced applications will require manipulators with redundant 'degrees of freedom' (DOF) that are capable of adapting manipulator configurations to avoid obstacles while achieving the user specified goal. Conventional methods for control of manipulators (based on solution of the inverse kinematics) cannot be easily extended to these situations. Fuzzy logic control offers a possible solution to these needs. A current research program at SRI developed a fuzzy logic controller for a redundant, 4 DOF, planar manipulator. The manipulator end point trajectory can be specified by either a computer program (robot mode) or by manual input (teleoperator). The approach used expresses end-point error and the location of manipulator joints as fuzzy variables. Joint motions are determined by a fuzzy rule set without requiring solution of the inverse kinematics. Additional rules for sensor data, obstacle avoidance and preferred manipulator configuration, e.g., 'righty' or 'lefty', are easily accommodated. The procedure used to generate the fuzzy rules can be extended to higher DOF systems.

  7. Relative Intensity Influences the Degree of Correspondence of Jump Squats and Push Jerks to Countermovement Jumps.

    PubMed

    Cushion, Emily J; Goodwin, Jon E; Cleather, Daniel J

    2016-05-01

    The aim of this study was to determine the mechanical similarity between push jerk (PJ) and jump squat (JS) to countermovement jump (CMJ) and further understand the effect increasing external load may have on this relationship. Eight physically trained men (age 22 ± 3; height 176 ± 7 kg; weight 83 ± 8 kg) performed an unloaded CMJ followed by JS under a range of loads (10, 25, 35, and 50% 1RM back squat) and PJ (30, 50, 65, and 75% 1RM push jerk). A portable force platform and high-speed camera both collecting at 250 Hz were used to establish joint moments and impulse during the propulsive phase of the movements. A standard inverse dynamics model was used to determine joint moment and impulse at the hip, knee, and ankle. Significant correlations (p ≤ 0.05) were shown between CMJ knee joint moment and JS knee joint moment at 25% load and PJ knee joint moment at 30 and 50% load. Significant correlations were also observed between CMJ knee joint impulse and JS knee joint impulse at 10% load and PJ knee joint moment at 30 and 65% load. Significant correlation was also observed between CMJ hip joint impulse and PJ hip joint impulse at 30% load. No significant joint × load interaction was shown as load increased for either PJ or JS. Results from the study suggest partial correspondence between PJ and JS to CMJ, where a greater mechanical similarity was observed between the PJ and CMJ. This interaction is load and joint dependent where lower relative loads showed greatest mechanical similarity. Therefore using lower relative loads when programming may provide a greater transfer of training effect.

  8. Learning to play the violin: motor control by freezing, not freeing degrees of freedom.

    PubMed

    Konczak, Jürgen; Vander Velden, Heidi; Jaeger, Lukas

    2009-05-01

    Playing a violin requires precise patterns of limb coordination that are acquired over years of practice. In the present study, the authors investigated how motion at proximal arm joints influenced the precision of bow movements in novice learners and experts. The authors evaluated the performances of 11 children (4-12 years old), 3 beginning-to-advanced level adult players, and 2 adult concert violinists, using a musical work that all had mastered as their first violin piece. The authors found that learning to play the violin was not associated with a release or freeing of joint degrees of freedom. Instead, learning was characterized by an experience-dependent suppression of sagittal shoulder motion, as documented by an observed reduction in joint angular amplitude. This reduction in the amplitude of shoulder flexion-extension correlated highly with a decrease of bow-movement variability. The remaining mechanical degrees of freedom at the elbow and shoulder showed patterns of neither suppression nor freeing. Only violinists with more than 700 practice hr achieved sagittal shoulder range of motion comparable to experts. The findings imply that restricting joint amplitude at selected joint degrees of freedom, while leaving other degrees of freedom unconstrained, constitutes an appropriate strategy for learning complex, high-precision motor patterns in children and adults. The findings also highlight that mastering even seemingly simple bowing movements constitutes a prolonged learning process.

  9. Improved Ball-and-Socket Docking Mechanism

    NASA Technical Reports Server (NTRS)

    Cloyd, Richard; Bryan, Tom

    2004-01-01

    A proposed docking mechanism would form a ball-and-socket joint in the docked condition. The mechanism could tolerate significant initial misalignment because it would include an alignment cone that would guide the ball into the socket. Like other ball-and-socket joints, the joint would have three rotational degrees of freedom. This docking mechanism would be a successor to the one described in Passive Capture Joint With Three Degrees of Freedom (MFS-31146), NASA Tech Briefs, Vol. 22, No. 7 (July 1998), page 65. It would contain most of the components of the prior mechanism, plus some additional components that would expand its capabilities.

  10. Six degree-of-freedom knee joint kinematics in obese individuals with knee pain during gait.

    PubMed

    Li, Jing-Sheng; Tsai, Tsung-Yuan; Felson, David T; Li, Guoan; Lewis, Cara L

    2017-01-01

    Knee joint pain is a common symptom in obese individuals and walking is often prescribed as part of management programs. Past studies in obese individuals have focused on standing alignment and kinematics in the sagittal and coronal planes. Investigation of 6 degree-of-freedom (6DOF) knee joint kinematics during standing and gait is important to thoroughly understand knee function in obese individuals with knee pain. This study aimed to investigate the 6DOF knee joint kinematics in standing and during gait in obese patients using a validated fluoroscopic imaging system. Ten individuals with obesity and knee pain were recruited. While standing, the knee was in 7.4±6.3°of hyperextension, 2.8±3.3° of abduction and 5.6±7.3° of external rotation. The femoral center was located 0.7±3.1mm anterior and 5.1±1.5mm medial to the tibial center. During treadmill gait, the sagittal plane motion, i.e., flexion/extension and anterior-posterior translation, showed a clear pattern. Specifically, obese individuals with knee pain maintained the knee in more flexion and more anterior tibial translation during most of the stance phase of the gait cycle and had a reduced total range of knee flexion when compared to a healthy non-obese group. In conclusion, obese individuals with knee pain used hyperextension knee posture while standing, but maintained the knee in more flexion during gait with reduced overall range of motion in the 6DOF analysis.

  11. Six degree-of-freedom knee joint kinematics in obese individuals with knee pain during gait

    PubMed Central

    Li, Jing-Sheng; Tsai, Tsung-Yuan; Felson, David T.; Li, Guoan; Lewis, Cara L.

    2017-01-01

    Knee joint pain is a common symptom in obese individuals and walking is often prescribed as part of management programs. Past studies in obese individuals have focused on standing alignment and kinematics in the sagittal and coronal planes. Investigation of 6 degree-of-freedom (6DOF) knee joint kinematics during standing and gait is important to thoroughly understand knee function in obese individuals with knee pain. This study aimed to investigate the 6DOF knee joint kinematics in standing and during gait in obese patients using a validated fluoroscopic imaging system. Ten individuals with obesity and knee pain were recruited. While standing, the knee was in 7.4±6.3°of hyperextension, 2.8±3.3° of abduction and 5.6±7.3° of external rotation. The femoral center was located 0.7±3.1mm anterior and 5.1±1.5mm medial to the tibial center. During treadmill gait, the sagittal plane motion, i.e., flexion/extension and anterior-posterior translation, showed a clear pattern. Specifically, obese individuals with knee pain maintained the knee in more flexion and more anterior tibial translation during most of the stance phase of the gait cycle and had a reduced total range of knee flexion when compared to a healthy non-obese group. In conclusion, obese individuals with knee pain used hyperextension knee posture while standing, but maintained the knee in more flexion during gait with reduced overall range of motion in the 6DOF analysis. PMID:28339477

  12. A 17 degree of freedom anthropomorphic manipulator

    NASA Technical Reports Server (NTRS)

    Vold, Havard I.; Karlen, James P.; Thompson, Jack M., Jr.; Farrell, James D.; Eismann, Paul H.

    1989-01-01

    A 17 axis anthropomorphic manipulator, providing coordinated control of two seven degree of freedom arms mounted on a three degree of freedom torso-waist assembly, is presented. This massively redundant telerobot, designated the Robotics Research K/B-2017 Dexterous Manipulator, employs a modular mechanism design with joint-mounted actuators based on brushless motors and harmonic drive gear reducers. Direct joint torque control at the servo level causes these high-output joint drives to behave like direct-drive actuators, facilitating the implementation of an effective impedance control scheme. The redundant, but conservative motion control system models the manipulator as a spring-loaded linkage with viscous damping and rotary inertia at each joint. This approach allows for real time, sensor-driven control of manipulator pose using a hierarchy of competing rules, or objective functions, to avoid unplanned collisions with objects in the workplace, to produce energy-efficient, graceful motion, to increase leverage, to control effective impedance at the tool or to favor overloaded joints.

  13. The effect of different decline angles on the biomechanics of double limb squats and the implications to clinical and training practice.

    PubMed

    Richards, Jim; Selfe, James; Sinclair, Jonathan; May, Karen; Thomas, Gavin

    2016-09-01

    Bilateral decline squatting has been well documented as a rehabilitation exercise, however, little information exists on the optimum angle of decline. The aim of this study was to determine the ankle and knee angle, moments, the patellofemoral joint load, patellar tendon load and associated muscle activity while performing a double limb squat at different decline angles and the implications to rehabilitation. Eighteen healthy subjects performed double limb squats at 6 angles of declination: 0, 5, 10, 15, 20 and 25 degrees. The range of motion of the knee and ankle joints, external moments, the patellofemoral/patellar tendon load and integrated EMG of gastrocnemius, tibialis anterior, rectus femoris and biceps femoris were evaluated. As the decline angle increased up to 20 degrees, the range of motion possible at the ankle and knee increased. The joint moments showed a decrease at the ankle up to 15 degrees and an increase at the knee up to 25 degrees, indicating a progressive reduction in loading around the ankle with a corresponding increase of the load in the patellar tendon and patellofemoral joint. These trends were supported by a decrease in tibialis anterior activity and an increase in the rectus femoris activity up to 15 degrees declination. However, gastrocnemius and biceps femoris activity increased as the decline angle increased above 15 degrees. The action of gastrocnemius and biceps femoris stabilises the knee against an anterior displacement of the femur on the tibia. These findings would suggest that there is little benefit in using a decline angle greater than 15-20 degrees unless the purpose is to offer an additional stability challenge to the knee joint.

  14. The effect of different decline angles on the biomechanics of double limb squats and the implications to clinical and training practice

    PubMed Central

    Richards, Jim; Selfe, James; Sinclair, Jonathan; May, Karen; Thomas, Gavin

    2016-01-01

    Abstract Bilateral decline squatting has been well documented as a rehabilitation exercise, however, little information exists on the optimum angle of decline. The aim of this study was to determine the ankle and knee angle, moments, the patellofemoral joint load, patellar tendon load and associated muscle activity while performing a double limb squat at different decline angles and the implications to rehabilitation. Eighteen healthy subjects performed double limb squats at 6 angles of declination: 0, 5, 10, 15, 20 and 25 degrees. The range of motion of the knee and ankle joints, external moments, the patellofemoral/patellar tendon load and integrated EMG of gastrocnemius, tibialis anterior, rectus femoris and biceps femoris were evaluated. As the decline angle increased up to 20 degrees, the range of motion possible at the ankle and knee increased. The joint moments showed a decrease at the ankle up to 15 degrees and an increase at the knee up to 25 degrees, indicating a progressive reduction in loading around the ankle with a corresponding increase of the load in the patellar tendon and patellofemoral joint. These trends were supported by a decrease in tibialis anterior activity and an increase in the rectus femoris activity up to 15 degrees declination. However, gastrocnemius and biceps femoris activity increased as the decline angle increased above 15 degrees. The action of gastrocnemius and biceps femoris stabilises the knee against an anterior displacement of the femur on the tibia. These findings would suggest that there is little benefit in using a decline angle greater than 15-20 degrees unless the purpose is to offer an additional stability challenge to the knee joint. PMID:28149400

  15. Prevalence and inheritance of and selection for elbow arthrosis in Bernese mountain dogs and Rottweilers in Sweden and benefit: cost analysis of a screening and control program.

    PubMed

    Swenson, L; Audell, L; Hedhammar, A

    1997-01-15

    To determine the prevalence and charges over time in the prevalence of elbow arthrosis in Bernese Mountain Dogs and Rottweilers, to ascertain whether prevalence or severity of elbow arthrosis was associated with sex of the dogs, age at the time of elbow joint examination, or ancestral background, to determine the effects of selective breeding, and to conduct an economic evaluation of the elbow arthrosis program operated by the Swedish Kennel Club. Analysis of radiographic evaluations of elbow joint conformity. 4,515 dogs from 2 breeds registered by the Swedish Kennel Club. All radiographs were scrutinized by a single radiologist (LA), and elbow joint conformation was classified as normal or arthrotic, with the degree of arthrosis classified as 1, 2, or 3. Decreasing prevalence of elbow arthrosis corresponding to selection of breeding stock and high heritabilities was found. Sex differences were documented in both breeds, but with contradictory directions. This was interpreted as breed differences in the distribution of genes related to elbow arthrosis. Economic analyses showed that costs of screening and registration of elbow joints was less than the value of dogs estimated to have been saved from moderate and severe elbow arthrosis in both breeds. Documented effects of age suggest that all dogs should be screened at the same age, rather than screening a few dogs at an older, more revealing age. In screening and control programs based on an open registry with access to family records, decreasing prevalence of elbow arthrosis can be expected, and related to selection of breeding stock.

  16. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Dwyer, Morgan Maeve

    This report summarizes the results of doctoral research that explored the cost impact of acquiring complex government systems jointly. The report begins by reviewing recent evidence that suggests that joint programs experience greater cost growth than non-joint programs. It continues by proposing an alternative approach for studying cost growth on government acquisition programs and demonstrates the utility of this approach by applying it to study the cost of jointness on three past programs that developed environmental monitoring systems for low-Earth orbit. Ultimately, the report concludes that joint programs' costs grow when the collaborating government agencies take action to retain ormore » regain their autonomy. The report provides detailed qualitative and quantitative data in support of this conclusion and generalizes its findings to other joint programs that were not explicitly studied here. Finally, it concludes by presenting a quantitative model that assesses the cost impacts of jointness and by demonstrating how government agencies can more effectively architect joint programs in the future.« less

  17. Meaningful Engagement to Enhance Diversity: Broadened Impact Actualized

    NASA Astrophysics Data System (ADS)

    Whitney, V. W.; Pyrtle, A. J.

    2008-12-01

    The MS PHD'S Professional Development Program was established by and for UR/US populations to facilitate increased and sustained participation within the Earth system science community. MS PHD'S is jointly funded by NSF and NASA. Fourteen (14) minority Earth system scientists served as Program mentors and one- hundred fifteen (115) minority and non-minority scientists served as Meeting Mentors to student participants. Representatives from fifty-six (56) agencies and institutions provided support and exposure to MS PHD'S student participants. Two hundred fifty-eight (258) highly qualified UR/US students completed on-line applications to participate in the MS PHD'S Professional Development Program. Because of funding limitations, slightly fewer than 50% of the applicants were selected to participate. One-hundred twenty-six (126) undergraduate and graduate students from 26 states and Puerto Rico participated in the MS PHD'S program. Sixty-eight (68) MS PHD'S student participants self-identified as African American; thirty-four (34) as Puerto Rican; nine (9) as Hispanic/Mexican American, ten (10) as Native American and one (1) each as African, Asian, Pacific Islander, Hispanic and Multi-Ethnic. During the five year span of MS PHD'S programming, sixteen (16) student participants completed BS degrees, twelve (12) completed MS degrees and ten (10) completed the Doctoral degrees. How did MS PHD'S establish meaningful engagement to enhance diversity within the Earth system science community? This case study reveals replicable processes and constructs to enhance the quality of meaningful collaboration and engagement. In addition, the study addresses frequently asked questions (FAQ's) on outreach, recruitment, engagement, retention and success of students from underrepresented populations within diversity-focused programs.

  18. Academic Responses to Fukushima Disaster.

    PubMed

    Yasui, Kiyotaka; Kimura, Yuko; Kamiya, Kenji; Miyatani, Rie; Tsuyama, Naohiro; Sakai, Akira; Yoshida, Koji; Yamashita, Shunichi; Chhem, Rethy; Abdel-Wahab, May; Ohtsuru, Akira

    2017-03-01

    Since radiation accidents, particularly nuclear disasters, are rarer than other types of disasters, a comprehensive radiation disaster medical curriculum for them is currently unavailable. The Fukushima compound disaster has urged the establishment of a new medical curriculum in preparation for any future complex disaster. The medical education will aim to aid decision making on various health risks for workers, vulnerable people, and residents addressing each phase in the disaster. Herein, we introduce 3 novel educational programs that have been initiated to provide students, professionals, and leaders with the knowledge of and skills to elude the social consequences of complex nuclear disasters. The first program concentrates on radiation disaster medicine for medical students at the Fukushima Medical University, together with a science, technology, and society module comprising various topics, such as public risk communication, psychosocial consequences of radiation anxiety, and decision making for radiation disaster. The second program is a Phoenix Leader PhD degree at the Hiroshima University, which aims to develop future leaders who can address the associated scientific, environmental, and social issues. The third program is a Joint Graduate School of Master's degree in the Division of Disaster and Radiation Medical Sciences at the Nagasaki University and Fukushima Medical University.

  19. Configuration control of seven-degree-of-freedom arms

    NASA Technical Reports Server (NTRS)

    Seraji, Homayoun (Inventor); Long, Mark K. (Inventor); Lee, Thomas S. (Inventor)

    1992-01-01

    A seven degree of freedom robot arm with a six degree of freedom end effector is controlled by a processor employing a 6 by 7 Jacobian matrix for defining location and orientation of the end effector in terms of the rotation angles of the joints, a 1 (or more) by 7 Jacobian matrix for defining 1 (or more) user specified kinematic functions constraining location or movement of selected portions of the arm in terms of the joint angles, the processor combining the two Jacobian matrices to produce an augmented 7 (or more) by 7 Jacobian matrix, the processor effecting control by computing in accordance with forward kinematics from the augmented 7 by 7 Jacobian matrix and from the seven joint angles of the arm a set of seven desired joint angles for transmittal to the joint servo loops of the arm. One of the kinematic functions constraints the orientation of the elbow plane of the arm. Another one of the kinematic functions minimizes a sum of gravitational torques on the joints. Still another kinematic function constrains the location of the arm to perform collision avoidance. Generically, one kinematic function minimizes a sum of selected mechanical parameters of at least some of the joints associated with weighting coefficients which may be changed during arm movement. The mechanical parameters may be velocity errors or gravity torques associated with individual joints.

  20. Configuration control of seven degree of freedom arms

    NASA Technical Reports Server (NTRS)

    Seraji, Homayoun (Inventor)

    1995-01-01

    A seven-degree-of-freedom robot arm with a six-degree-of-freedom end effector is controlled by a processor employing a 6-by-7 Jacobian matrix for defining location and orientation of the end effector in terms of the rotation angles of the joints, a 1 (or more)-by-7 Jacobian matrix for defining 1 (or more) user-specified kinematic functions constraining location or movement of selected portions of the arm in terms of the joint angles, the processor combining the two Jacobian matrices to produce an augmented 7 (or more)-by-7 Jacobian matrix, the processor effecting control by computing in accordance with forward kinematics from the augmented 7-by-7 Jacobian matrix and from the seven joint angles of the arm a set of seven desired joint angles for transmittal to the joint servo loops of the arms. One of the kinematic functions constrains the orientation of the elbow plane of the arm. Another one of the kinematic functions minimizing a sum of gravitational torques on the joints. Still another one of the kinematic functions constrains the location of the arm to perform collision avoidance. Generically, one of the kinematic functions minimizes a sum of selected mechanical parameters of at least some of the joints associated with weighting coefficients which may be changed during arm movement. The mechanical parameters may be velocity errors or position errors or gravity torques associated with individual joints.

  1. Development and initial evaluation of a training program for peripherally inserted central catheter (PICC) placement for radiology residents and technicians.

    PubMed

    Dabadie, A; Soussan, J; Mancini, J; Vidal, V; Bartoli, J M; Gorincour, G; Petit, P

    2016-09-01

    The goals of this study were to develop and evaluate a joint theoretical/practical training course for radiology residents and technicians and to start a collaborative practice agreement enabling radiology technicians to perform PICC placement under the responsibility of an interventional radiologist. A joint training session based on literature evidences and international recommendations was designed. Participants were assessed before and after training, and were also asked to evaluate the program one month after completion of the training course. Practical post-training mentoring guidelines were laid down for radiologists supervising technicians. From January to April 2014, 6 radiology residents and 12 radiology technicians from the two interventional radiology departments of the University hospitals in Marseille took part in the training program. For both residents and technicians, significant improvement was observed between pretraining and post-training assessment. The majority of participants were satisfied with the program. Our experience suggests that combined theoretical and practical training in PICC placement allows improving technical skill and yields high degrees of satisfaction for both radiology residents and technicians. A collaborative practice agreement is now formally established to enable radiologists to delegate PICC placement procedures to radiology technicians. Copyright © 2016 Éditions françaises de radiologie. Published by Elsevier Masson SAS. All rights reserved.

  2. Electrical stimulation on joint contracture: an experiment in rat model with direct current.

    PubMed

    Akai, M; Shirasaki, Y; Tateishi, T

    1997-04-01

    To examine whether electrical stimulation could decrease the degree of joint stiffness in a rat lower extremity model. Rat knee joints were surgically immobilized in a flexed position for 3 weeks. Two groups of rats were stimulated with 20 microA and 50 microA constant direct current. Another group had surgical intervention and sham electrodes without electricity. The hind leg was extirpated and prepared for a sample with the femur-knee joint-tibia unit. Recording the knee flexion angle with extension torque, the degree of joint contracture was assessed biomechanically by measuring the bone-joint-bone sample as a cantilever. Measurement was performed with (1) spectral analysis of transfer function measurement using random mechanical noise with frequency range from 1 to 50Hz, and (2) dynamic stiffness and loss tangent with steady-state sinusoidal excitation (11 and 35Hz). The results showed that no significant difference or trend was found in vibration analysis among three groups. However, spectral analysis of transfer function measurement revealed more deformation against load, and more viscous nature in the stimulation groups, especially in low frequency band, than in the sham group. Electrical stimulation with constant direct current has a possibility of reducing the degree of joint contracture.

  3. The influence of patellofemoral joint contact geometry on the modeling of three dimensional patellofemoral joint forces.

    PubMed

    Powers, Christopher M; Chen, Yu-Jen; Scher, Irving; Lee, Thay Q

    2006-01-01

    The purpose of this study was to determine the influence of patellofemoral joint contact geometry on the modeling of three-dimensional patellofemoral joint forces. To achieve this goal, patellofemoral joint reaction forces (PFJRFs) that were measured from an in-vitro cadaveric set-up were compared to PFJRFs estimated from a computer model that did not consider patellofemoral joint contact geometry. Ten cadaver knees were used in this study. Each was mounted on a custom jig that was fixed to an Instron frame. Quadriceps muscle loads were accomplished using a pulley system and weights. The force in the patellar ligament was obtained using a buckle transducer. To quantify the magnitude and direction of the PFJRF, a six-axis load cell was incorporated into the femoral fixation system so that a rigid body assumption could be made. PFJRF data were obtained at 0 degrees , 20 degrees , 40 degrees and 60 degrees of knee flexion. Following in vitro testing, SIMM modeling software was used to develop computational models based on the three-dimensional coordinates (Microscribe digitizer) of individual muscle and patellar ligament force vectors obtained from the cadaver knees. The overall magnitude of the PFJRF estimated from the computer generated models closely matched the direct measurements from the in vitro set-up (Pearson's correlation coefficient, R(2)=0.91, p<0.001). Although the computational model accurately estimated the posteriorly directed forces acting on the joint, some discrepancies were noted in the forces acting in the superior and lateral directions. These differences however, were relatively small when expressed as a total of the overall PFJRF magnitude.

  4. Joint Task Force on Undergraduate Physics Programs (J-TUPP): Overview and Major Findings

    NASA Astrophysics Data System (ADS)

    Heron, Paula

    2016-03-01

    The Joint Task Force on Undergraduate Physics Programs (JTUPP) was formed in response to growing awareness in the physics community that physics majors pursue a wide range of careers after graduation, with very few ending up in academia. The task force is charged with identifying the skills and knowledge that undergraduate physics degree holders should possess to be well prepared for a diverse set of careers, and providing guidance for physicists considering revising the undergraduate curriculum to improve the education of a diverse student population. Task force members represent large and small universities, professional societies, and industry, and have expertise in a broad range of areas including entrepreneurship, physics education research and systemic change in education. We reviewed employment data, surveys of employers, and reports generated by other disciplines. We also met with physicists in selected industries to get their views on the strengths and weaknesses of physics graduates, commissioned a series of interviews with recent physics graduates employed in the private sector, and identified exemplary programs that ensure that all of their students are well prepared to pursue a wide range of career paths. The findings and recommendations will be summarized.

  5. Design of a Minimum Surface-Effect Three Degree-of-Freedom Micromanipulator

    NASA Technical Reports Server (NTRS)

    Goldfarb, Michael; Speich, John E.

    1997-01-01

    This paper describes the fundamental physical motivations for small-scale minimum surface-effect design, and presents a three degree-of-freedom micromanipulator design that incorporates a minimum surface-effect approach. The primary focus of the design is the split-tube flexure, a unique small-scale revolute joint that exhibits a considerably larger range of motion and significantly better multi-axis revolute joint characteristics than a conventional flexure. The development of this joint enables the implementation of a small-scale spatially-loaded revolute joint-based manipulator with well-behaved kinematic characteristics and without the backlash and stick-slip behavior that would otherwise prevent precision control

  6. Jerk-level synchronous repetitive motion scheme with gradient-type and zeroing-type dynamics algorithms applied to dual-arm redundant robot system control

    NASA Astrophysics Data System (ADS)

    Chen, Dechao; Zhang, Yunong

    2017-10-01

    Dual-arm redundant robot systems are usually required to handle primary tasks, repetitively and synchronously in practical applications. In this paper, a jerk-level synchronous repetitive motion scheme is proposed to remedy the joint-angle drift phenomenon and achieve the synchronous control of a dual-arm redundant robot system. The proposed scheme is novelly resolved at jerk level, which makes the joint variables, i.e. joint angles, joint velocities and joint accelerations, smooth and bounded. In addition, two types of dynamics algorithms, i.e. gradient-type (G-type) and zeroing-type (Z-type) dynamics algorithms, for the design of repetitive motion variable vectors, are presented in detail with the corresponding circuit schematics. Subsequently, the proposed scheme is reformulated as two dynamical quadratic programs (DQPs) and further integrated into a unified DQP (UDQP) for the synchronous control of a dual-arm robot system. The optimal solution of the UDQP is found by the piecewise-linear projection equation neural network. Moreover, simulations and comparisons based on a six-degrees-of-freedom planar dual-arm redundant robot system substantiate the operation effectiveness and tracking accuracy of the robot system with the proposed scheme for repetitive motion and synchronous control.

  7. PA03.05. Masha taila as a preventive measure in gulpha marma injuries.

    PubMed Central

    Dhaded, Rajani; Kulkarni, BG

    2013-01-01

    Purpose: In Basketball game Gulpha Marma injuries are common. Wrong landing causes inflammation & tearing of ligament resulting into sprain. Prevention can be carried out by protective wears like Ankelet,crape bandge which gives temporary recovery. Assessment and evaluation of the application of Masha Taila as a Snehana and Avagaha Sweda as a preventive measure for Gulpha Marma injuries. Method: 30 players with Normal ankle joint (Group A) & 30 players with Ankle sprain (1st degree) of Basketball players (Group B) were taken with age group 13 to 25 yrs. Goniometer to measure Range of motion (R.O.M.), pain analog scale for pain gradation. Result: Group A and B were compared with their control groups. The statistical result for normal dorsiflexion was 0.0007 degree, plantar flexion 0.001 degree, dorsi flexion of ankle sprain players 0.002 degree, plantar flexion 0.03 degree. The measurement of ROM of all players increased by 510 times indicating increased joint flexibility, based on ROM Mash taila proved to be a preventive measure. Pain analog scale for group B players indicated moderate pain at 0 day,reduced by 50% on 15th day and completely reduced on 30th day. Conclusion: “Prevention is better than cure”. As a preventive aspect the application of Masha Taila in normal ankle joint as Snehan & Swedan with warm water increases flexibility of joint and muscle strength. In sprained joint the same Taila application relives pain and regains its movements, this recovery and strength of joints happens because of reduce adhesion & influence the direction of new collage fibres in the healing process. It eliminates toxic accumulation from secondary muscle spasm. These Marma being the vital points must be prevented by applying Snehan & Swedan before starting the game just as warm up.

  8. Double slotted socket spherical joint

    DOEpatents

    Bieg, Lothar F.; Benavides, Gilbert L.

    2001-05-22

    A new class of spherical joints is disclosed. These spherical joints are capable of extremely large angular displacements (full cone angles in excess of 270.degree.), while exhibiting no singularities or dead spots in their range of motion. These joints can improve or simplify a wide range of mechanical devices.

  9. Increasing Ti-6Al-4V brazed joint strength equal to the base metal by Ti and Zr amorphous filler alloys

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ganjeh, E., E-mail: navidganjehie@sina.kntu.ac.ir; Sarkhosh, H.; Bajgholi, M.E.

    Microstructural features developed along with mechanical properties in furnace brazing of Ti-6Al-4V alloy using STEMET 1228 (Ti-26.8Zr-13Ni-13.9Cu, wt.%) and STEMET 1406 (Zr-9.7Ti-12.4Ni-11.2Cu, wt.%) amorphous filler alloys. Brazing temperatures employed were 900-950 Degree-Sign C for the titanium-based filler and 900-990 Degree-Sign C for the zirconium-based filler alloys, respectively. The brazing time durations were 600, 1200 and 1800 s. The brazed joints were evaluated by ultrasonic test, and their microstructures and phase constitutions analyzed by metallography, scanning electron microscopy and X-ray diffraction analysis. Since microstructural evolution across the furnace brazed joints primarily depends on their alloying elements such as Cu, Ni andmore » Zr along the joint. Accordingly, existence of Zr{sub 2}Cu, Ti{sub 2}Cu and (Ti,Zr){sub 2}Ni intermetallic compounds was identified in the brazed joints. The chemical composition of segregation region in the center of brazed joints was identical to virgin filler alloy content which greatly deteriorated the shear strength of the joints. Adequate brazing time (1800 s) and/or temperature (950 Degree-Sign C for Ti-based and 990 Degree-Sign C for Zr-based) resulted in an acicular Widmanstaetten microstructure throughout the entire joint section due to eutectoid reaction. This microstructure increased the shear strength of the brazed joints up to the Ti-6Al-4V tensile strength level. Consequently, Ti-6Al-4V can be furnace brazed by Ti and Zr base foils produced excellent joint strengths. - Highlights: Black-Right-Pointing-Pointer Temperature or time was the main factors of controlling braze joint strength. Black-Right-Pointing-Pointer Developing a Widmanstaetten microstructure generates equal strength to base metal. Black-Right-Pointing-Pointer Brittle intermetallic compounds like (Ti,Zr){sub 2}Ni/Cu deteriorate shear strength. Black-Right-Pointing-Pointer Ti and Zr base filler alloys were the best choice for brazing Ti-6Al-4V.« less

  10. Analysis of Knee Joint Line Obliquity after High Tibial Osteotomy.

    PubMed

    Oh, Kwang-Jun; Ko, Young Bong; Bae, Ji Hoon; Yoon, Suk Tae; Kim, Jae Gyoon

    2016-11-01

    The aim of this study was to evaluate which lower extremity alignment (knee and ankle joint) parameters affect knee joint line obliquity (KJLO) in the coronal plane after open wedge high tibial osteotomy (OWHTO). Overall, 69 knees of patients that underwent OWHTO were evaluated using radiographs obtained preoperatively and from 6 weeks to 3 months postoperatively. We measured multiple parameters of knee and ankle joint alignment (hip-knee-ankle angle [HKA], joint line height [JLH], posterior tibial slope [PS], femoral condyle-tibial plateau angle [FCTP], medial proximal tibial angle [MPTA], mechanical lateral distal femoral angle [mLDFA], KJLO, talar tilt angle [TTA], ankle joint obliquity [AJO], and the lateral distal tibial ground surface angle [LDTGA]; preoperative [-pre], postoperative [-post], and the difference between -pre and -post values [-Δ]). We categorized patients into two groups according to the KJLO-post value (the normal group [within ± 4 degrees, 56 knees] and the abnormal group [greater than ± 4 degrees, 13 knees]), and compared their -pre parameters. Multiple logistic regression analysis was used to examine the contribution of the -pre parameters to abnormal KJLO-post. The mean HKA-Δ (-9.4 ± 4.7 degrees) was larger than the mean KJLO-Δ (-2.1 ± 3.2 degrees). The knee joint alignment parameters (the HKA-pre, FCTP-pre) differed significantly between the two groups ( p  < 0.05). In addition, the HKA-pre (odds ratio [OR] = 1.27, p  = 0.006) and FCTP-pre (OR = 2.13, p  = 0.006) were significant predictors of abnormal KJLO-post. However, -pre ankle joint parameters (TTA, AJO, and LDTGA) did not differ significantly between the two groups and were not significantly associated with the abnormal KJLO-post. The -pre knee joint alignment and knee joint convergence angle evaluated by HKA-pre and FCTP-pre angle, respectively, were significant predictors of abnormal KJLO after OWHTO. However, -pre ankle joint parameters were not significantly associated with abnormal KJLO after OWHTO. Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

  11. Mechanical stability of the subtalar joint after lateral ligament sectioning and ankle brace application: a biomechanical experimental study.

    PubMed

    Kamiya, Tomoaki; Kura, Hideji; Suzuki, Daisuke; Uchiyama, Eiichi; Fujimiya, Mineko; Yamashita, Toshihiko

    2009-12-01

    The roles of each ligament supporting the subtalar joint have not been clarified despite several biomechanical studies. The effects of ankle braces on subtalar instability have not been shown. The ankle brace has a partial effect on restricting excessive motion of the subtalar joint. Controlled laboratory study. Ten normal fresh-frozen cadaveric specimens were used. The angular motions of the talus were measured via a magnetic tracking system. The specimens were tested while inversion and eversion forces, as well as internal and external rotation torques, were applied. The calcaneofibular ligament, cervical ligament, and interosseous talocalcaneal ligament were sectioned sequentially, and the roles of each ligament, as well as the stabilizing effects of the ankle brace, were examined. Complete sectioning of the ligaments increased the angle between the talus and calcaneus in the frontal plane to 51.7 degrees + or - 11.8 degrees compared with 35.7 degrees + or - 6.0 degrees in the intact state when inversion force was applied. There was a statistically significant difference in the angles between complete sectioning of the ligaments and after application of the brace (34.1 degrees + or - 7.3 degrees ) when inversion force was applied. On the other hand, significant differences in subtalar rotation were not found between complete sectioning of the ligaments and application of the brace when internal and external rotational torques were applied. The ankle brace limited inversion of the subtalar joint, but it did not restrict motion after application of internal or external rotational torques. In cases of severe ankle sprains involving the calcaneofibular ligament, cervical ligament, and interosseous talocalcaneal ligament injuries, application of an ankle brace might be less effective in limiting internal-external rotational instabilities than in cases of inversion instabilities in the subtalar joint. An improvement in the design of the brace is needed to restore better rotational stability in the subtalar joint.

  12. Program Responses to Acute and Chronic Malnutrition: Divergences and Convergences123

    PubMed Central

    Bergeron, Gilles; Castleman, Tony

    2012-01-01

    Program approaches for addressing acute malnutrition and those for addressing chronic malnutrition have grown in different directions. Their specialization has led to productive advances in the efficacy of specific interventions but has also created divergences in implementation. Greater convergence and integration between the 2 sets of approaches would help programs respond to the diversity of conditions faced in the field and enable a more comprehensive continuum of care from prevention to treatment. After reviewing the causes of the differences in approach, this paper examines programmatic and scientific challenges to greater convergence and suggests steps to promote effective integration of acute and chronic malnutrition services. Steps include strengthening community linkages between program platforms, assessing the degree and type of integration needed in various situations, identifying cost efficiencies, and developing joint tools where possible. PMID:22516735

  13. A Program of Research and Education in Aerospace Structures at the Joint Institute for Advancement of Flight Sciences

    NASA Technical Reports Server (NTRS)

    Tolson, Robert H.

    2000-01-01

    The objectives of the cooperative effort with NASA was to conduct research related to aerospace structures and to increase the quality and quantity of highly trained engineers knowledgeable about aerospace structures. The program has successfully met the objectives and has been of significant benefit to NASA LARC, the GWU and the nation. The program was initiated with 3 students in 1994 under the direction of Dr. Robert Tolson as the Principal Investigator. Since initiation, 14 students have been involved in the program, resulting in 11 MS degrees with 2 more expected in 2000. The 11 MS theses and projects are listed. For technology transfer purposes some research is not reported in thesis form. Graduates from the program have been hired at aerospace and other companies across the nation, providing GWU and LARC with important industry and government contacts.

  14. Multi-planar bending properties of lumbar intervertebral joints following cyclic bending.

    PubMed

    Chow, Daniel H K; Luk, Keith D K; Holmes, Andrew D; Li, Xing-Fei; Tam, Steven C W

    2004-02-01

    To assess the changes in the multi-planar bending properties of intervertebral joints following cyclic bending along different directions. An in vitro biomechanical study using porcine lumbar motion segments. Repeated bending has been suggested as part of the etiology of gradual prolapse of the intervertebral disc, but the multi-planar changes in bending properties following cyclic loading have not been examined in detail. Porcine lumbar motion segments were subject to 1500 cycles of bending along directions of 0 degrees (flexion), 30 degrees, 60 degrees, or 90 degrees (right lateral bending). The multi-planar bending moments and hysteresis energies were recorded before loading and after various cycle numbers. Repeated bending at 30 degrees and 60 degrees resulted in greater decreases in mean bending moment and hysteresis energy than bending at 0 degrees or 90 degrees. No significant differences were seen between loading groups for the change in bending moment along the anterior testing directions, but significant differences were observed in the posterior and lateral testing directions, with bending at 30 degrees causing a significantly greater decrease in bending moment in the postero-lateral directions. The change in mechanical properties of porcine intervertebral joints due to cyclic bending depend on the direction of loading and the direction in which the properties are measured. Loading at 30 degrees provokes the most marked changes in bending moment and hysteresis energy.

  15. Joining of alumina via copper/niobium/copper interlayers

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Marks, Robert A.; Chapman, Daniel R.; Danielson, David T.

    2000-03-15

    Alumina has been joined at 1150 degrees C and 1400 degrees C using multilayer copper/niobium/copper interlayers. Four-point bend strengths are sensitive to processing temperature, bonding pressure, and furnace environment (ambient oxygen partial pressure). Under optimum conditions, joints with reproducibly high room temperature strengths (approximately equal 240 plus/minus 20 MPa) can be produced; most failures occur within the ceramic. Joints made with sapphire show that during bonding an initially continuous copper film undergoes a morphological instability, resulting in the formation of isolated copper-rich droplets/particles at the sapphire/interlayer interface, and extensive regions of direct bonding between sapphire and niobium. For optimized aluminamore » bonds, bend tests at 800 degrees C-1100 degrees C indicate significant strength is retained; even at the highest test temperature, ceramic failure is observed. Post-bonding anneals at 1000 degrees C in vacuum or in gettered argon were used to assess joint stability and to probe the effect of ambient oxygen partial pressure on joint characteristics. Annealing in vacuum for up to 200 h causes no significant decrease in room temperature bend strength or change in fracture path. With increasing anneal time in a lower oxygen partial pressure environment, the fracture strength decreases only slightly, but the fracture path shifts from the ceramic to the interface.« less

  16. Vascularised knee joint transplantation in man: the first two years experience.

    PubMed

    Kirschner, M H; Brauns, L; Gonschorek, O; Bühren, V; Hofmann, G O

    2000-04-01

    To describe our early experience with a new technique for restoring destroyed knee joints to give reasonable functional results. Observational clinical trial. Level-1-Trauma centre, Germany. 5 patients with large bone defects of the knee and loss of the extensor apparatus caused either by serious injury alone, or infection after serious injury. Transplantation of fresh and perfused knee joints with a vascular pedicle from multiorgan donors under immunosuppression. Ability to walk, need to remove one transplanted joint. Four patients are able to walk, the range of movement being from 50 degrees-120 degrees. The first patient additionally had to be provided with a total knee joint arthroplasty. In the third patient the graft became infected and had to be removed. She finally had an arthrodesis and bone lengthening by the Ilizarov technique. Transplantation of the knee joint may be an alternative to bone lengthening or amputation for patients with total loss of the extensor apparatus.

  17. [Case-control study on bone setting manipulation for the treatment of over degree II supination-eversion fractures of ankle joint].

    PubMed

    Qi, Yue-Feng; Chen, Fa-Lin; Bao, Shu-Ren; Li, Cheng-Huan; Zhao, Xing-Wei; Liu, Shi-Ming; Chen, Wen-Xue; Li, Ye; Wang, Peng

    2012-08-01

    To explore therapeutic effects of bone setting manipulation for the treatment of over degree II supination-eversion fractures of ankle,and analyze manipulative reduction mechanism. From 2005 to 2008, 95 patients with over degree II supination-eversion fractures of ankle were treated respectively by manipulation and operation. There were 43 cases [11 males and 32 females with an average age of (44.95 +/- 12.65) years] in manipulation group, and 2 cases were degree II, 11 cases were degree III, and 30 cases were degree IV. There were 52 cases [21 males and 31 females with an average age of (39.96 +/- 13.28) years] in operative group,and 6 cases were degree II, 18 cases were degree III, and 28 cases were degree IV. Bone setting manipulation and hard splint external fixation were applied to manipulative group. Operative reduction internal fixation was performed in operative group. X-ray was used to evaluate reduction of fracture before and after treatment, 2 months after treatment. Ankle joint function was evaluated according to Olerud-Molander scoring system after 6 months treatment. All patients were followed up with good reduction. Three cases occurred wound complication in operative group, but not in manipulative group. In manipulation group, 19 cases got excellent results, 20 cases good and 4 cases fair; while in operative group, 30 cases got excellent results, 20 cases good and 2 cases poor. There were no significant differences in fracture reduction and ankle joint function recovery between two groups (P > 0.05). Efficacy of operative treatment was better than that of manipulative treatment at degree IV fracture (P < 0.05). Bone setting manipulation is a good method for treating supination-eversion ankle joint fractures, which has advantages of simple and safe operation, reliable efficacy. For ankle join fracture at degree IV, manipulative reduction should be adopted earlier, and operative treatment also necessary

  18. Biomechanical consequences of a tear of the posterior root of the medial meniscus. Surgical technique.

    PubMed

    Harner, Christopher D; Mauro, Craig S; Lesniak, Bryson P; Romanowski, James R

    2009-10-01

    Tears of the posterior root of the medial meniscus are becoming increasingly recognized. They can cause rapidly progressive arthritis, yet their biomechanical effects are not understood. The goal of this study was to determine the effects of posterior root tears of the medial meniscus and their repairs on tibiofemoral joint contact pressure and kinematics. Nine fresh-frozen cadaver knees were used. An axial load of 1000 N was applied with a custom testing jig at each of four knee-flexion angles: 0 degrees , 30 degrees , 60 degrees , and 90 degrees . The knees were otherwise unconstrained. Four conditions were tested: (1) intact, (2) a posterior root tear of the medial meniscus, (3) a repaired posterior root tear, and (4) a total medial meniscectomy. Fuji pressure-sensitive film was used to record the contact pressure and area for each testing condition. Kinematic data were obtained by using a robotic arm to record the position of the knees for each loading condition. Three-dimensional knee kinematics were analyzed with custom programs with use of previously described transformations. The measured variables were axial rotation, varus angulation, lateral translation, and anterior translation. In the medial compartment, a posterior root tear of the medial meniscus caused a 25% increase in peak contact pressure compared with that found in the intact condition (p < 0.001). Repair restored the peak contact pressure to normal. No difference was detected between the peak contact pressure after the total medial meniscectomy and that associated with the root tear. The peak contact pressure in the lateral compartment after the total medial meniscectomy was up to 13% greater than that for all other conditions (p = 0.026). Significant increases in external rotation and lateral tibial translation, compared with the values in the intact knee, were observed in association with the posterior root tear (2.98 degrees and 0.84 mm, respectively) and the meniscectomy (4.45 degrees and 0.80 mm, respectively), and these increases were corrected by the repair. This study demonstrated significant changes in contact pressure and knee joint kinematics due to a posterior root tear of the medial meniscus. Root repair was successful in restoring joint biomechanics to within normal conditions.

  19. Biomechanical consequences of a tear of the posterior root of the medial meniscus. Similar to total meniscectomy.

    PubMed

    Allaire, Robert; Muriuki, Muturi; Gilbertson, Lars; Harner, Christopher D

    2008-09-01

    Tears of the posterior root of the medial meniscus are becoming increasingly recognized. They can cause rapidly progressive arthritis, yet their biomechanical effects are not understood. The goal of this study was to determine the effects of posterior root tears of the medial meniscus and their repairs on tibiofemoral joint contact pressure and kinematics. Nine fresh-frozen cadaver knees were used. An axial load of 1000 N was applied with a custom testing jig at each of four knee-flexion angles: 0 degrees, 30 degrees, 60 degrees, and 90 degrees. The knees were otherwise unconstrained. Four conditions were tested: (1) intact, (2) a posterior root tear of the medial meniscus, (3) a repaired posterior root tear, and (4) a total medial meniscectomy. Fuji pressure-sensitive film was used to record the contact pressure and area for each testing condition. Kinematic data were obtained by using a robotic arm to record the position of the knees for each loading condition. Three-dimensional knee kinematics were analyzed with custom programs with use of previously described transformations. The measured variables were axial rotation, varus angulation, lateral translation, and anterior translation. In the medial compartment, a posterior root tear of the medial meniscus caused a 25% increase in peak contact pressure compared with that found in the intact condition (p < 0.001). Repair restored the peak contact pressure to normal. No difference was detected between the peak contact pressure after the total medial meniscectomy and that associated with the root tear. The peak contact pressure in the lateral compartment after the total medial meniscectomy was up to 13% greater than that for all other conditions (p = 0.026). Significant increases in external rotation and lateral tibial translation, compared with the values in the intact knee, were observed in association with the posterior root tear (2.98 degrees and 0.84 mm, respectively) and the meniscectomy (4.45 degrees and 0.80 mm, respectively), and these increases were corrected by the repair. This study demonstrated significant changes in contact pressure and knee joint kinematics due to a posterior root tear of the medial meniscus. Root repair was successful in restoring joint biomechanics to within normal conditions.

  20. [COMPARISON OF FEMORAL CONDYLAR TWIST ANGLE IN THREE DIMENSIONAL RECONSTRUCTION DIGITAL MODELS OF KNEE JOINT BASED ON TWO DIMENSIONAL IMAGES OF MRI AND CT].

    PubMed

    Huang, Zan; Li, Yanlin; Hu, Meng; Li, Jian; You, Zhimin; Wang, Guoliang; He, Chuan

    2015-02-01

    To study the difference of femoral condylar twist angle (CTA) measurement in three dimensional (3-D) reconstruction digital models of human knee joint based on the two dimensional (2-D) images of MRI and CT so as to provide a reference for selecting the best method of CTA measurement in preoperative design for the femoral prosthesis rotational position. The CTA of 10 human cadaveric knee joint was measured in 3-D digital models based on MRI (group A), in 3-D digital models based on CT (group B), in the cadaveric knee joint with cartilage (group C), and in the cadaveric knee joint without cartilage (group D), respectively. The statistical analysis of the differences was made among the measurements of the CTA. The CTA values measured in 3-D digital models were (6.43 ± 0.53) degrees in group A and (3.31 ± 1.07) degrees in group B, showing significant difference (t = 10.235, P = 0.000). The CTA values measured in the cadaveric knee joint were (5.21 ± 1.28) degrees in group C and (3.33 ± 1.12) degrees in group D, showing significant difference (t = 5.770, P = 0.000). There was significant difference in the CTA values between group B and group C (t = 5.779, P = 0.000), but no significant difference was found between group A and group C (t = 3.219, P = 0.110). The CTA values measured in the 3-D digital models based on MRI are closer to the actual values measured in the knee joint with cartilage, and benefit for preoperative plan.

  1. One-degree-of-freedom spherical model for the passive motion of the human ankle joint.

    PubMed

    Sancisi, Nicola; Baldisserri, Benedetta; Parenti-Castelli, Vincenzo; Belvedere, Claudio; Leardini, Alberto

    2014-04-01

    Mathematical modelling of mobility at the human ankle joint is essential for prosthetics and orthotic design. The scope of this study is to show that the ankle joint passive motion can be represented by a one-degree-of-freedom spherical motion. Moreover, this motion is modelled by a one-degree-of-freedom spherical parallel mechanism model, and the optimal pivot-point position is determined. Passive motion and anatomical data were taken from in vitro experiments in nine lower limb specimens. For each of these, a spherical mechanism, including the tibiofibular and talocalcaneal segments connected by a spherical pair and by the calcaneofibular and tibiocalcaneal ligament links, was defined from the corresponding experimental kinematics and geometry. An iterative procedure was used to optimize the geometry of the model, able to predict original experimental motion. The results of the simulations showed a good replication of the original natural motion, despite the numerous model assumptions and simplifications, with mean differences between experiments and predictions smaller than 1.3 mm (average 0.33 mm) for the three joint position components and smaller than 0.7° (average 0.32°) for the two out-of-sagittal plane rotations, once plotted versus the full flexion arc. The relevant pivot-point position after model optimization was found within the tibial mortise, but not exactly in a central location. The present combined experimental and modelling analysis of passive motion at the human ankle joint shows that a one degree-of-freedom spherical mechanism predicts well what is observed in real joints, although its computational complexity is comparable to the standard hinge joint model.

  2. Anatomical reconstruction of the lateral ligaments of the ankle with a gracilis autograft: a new technique using an interference fit anchoring system.

    PubMed

    Takao, Masato; Oae, Kazunori; Uchio, Yuji; Ochi, Mitsuo; Yamamoto, Haruyasu

    2005-06-01

    Few anatomical and minor invasive procedures have been reported for surgical reconstruction of the lateral ligaments to treat lateral instability of the ankle. Furthermore, there are no standards according to which ligaments should be reconstructed. A new technique for anatomically reconstructing the lateral ligaments of the ankle using an interference fit anchoring system and determining which ligaments need to be reconstructed according to the results of standard stress radiography of the talocrural and subtalar joints will be effective for treating lateral instability of the ankle. Case series; level of evidence, 4. Twenty-one patients with lateral instability of the ankle underwent surgery using the proposed interference fit anchoring system. Standard stress radiographs of the subtalar joint were performed, and if the talocalcaneal angle was less than 10 degrees , only the anterior talofibular ligament was reconstructed; if there was a 10 degrees or greater opening of the talocalcaneal angle, both the anterior talofibular ligament and the calcaneofibular ligament were reconstructed. In the 17 patients who received only the anterior talofibular ligament reconstruction, the mean talar tilt angle on standard stress radiography of the talocrural joint was 14.5 degrees +/- 1.7 degrees before surgery and 2.6 degrees +/- 0.8 degrees 2 years after surgery (P < .0001). For the 4 patients who had both the anterior talofibular ligament and calcaneofibular ligament reconstructed, the mean talar tilt angle was 16.5 degrees +/- 1.5 degrees before surgery and 3.0 degrees +/- 0.5 degrees 2 years after surgery (P = .0015). The overall mean talocalcaneal angle on standard stress radiography of the subtalar joint was 11.3 degrees +/- 1.4 degrees before surgery and 3.5 degrees +/- 0.8 degrees 2 years after surgery (P = .0060). The proposed system has several advantages, including anatomical reconstruction with normal stability and range of motion restored, the need for only a small incision during the reconstruction, and sufficient strength at the tendon graft-bone tunnel junction, in comparison with the tension strength of the lateral ligaments of the ankle.

  3. Respiratory Therapy Faculty Knowledge of and Attitudes Toward Interprofessional Education.

    PubMed

    Vernon, Marlo M; Moore, Nicole M; Cummins, Lisa-Anne; Reyes, Stephanie E; Mazzoli, Andrew J; Heboyan, Vahe; De Leo, Gianluca

    2017-07-01

    Interprofessional education (IPE) improves collaboration and patient care through joint education between health professions. Respiratory therapy (RT) faculty were surveyed to evaluate their knowledge and attitudes toward IPE. We report current opportunities for IPE from faculty and compare responses from associate's, bachelor's, and master's degree programs and profit versus nonprofit institutions. We developed an online survey based on IPE literature and questions modified for the RT discipline. The survey was distributed by email to 874 faculty from the Commission on Accreditation for Respiratory Care accredited programs. The response rate was 33%. Faculty identified IPE as an important component of RT education ( n = 207, 80%) but reported challenges in integrating IPE into current curriculum. Overall, communication was ranked as the most important IPE competency ( n = 104, 39%) and ethics least important ( n = 131, 49%). When asked how many credit hours are required to teach IPE, 48% of respondents reported that they were unsure of an appropriate time requirement. Significant differences between associate's and bachelor's/master's degree program faculty were found on the following topics: institutional resources needed for IPE ( P < .001), faculty availability ( P < .001), curriculum availability for IPE ( P = .02), and importance of including IPE at academic health center campuses ( P < .001). IPE is recognized as an important component of RT education by all faculty respondents. However, significant differences in knowledge and attitudes toward IPE exist between faculty in associate's versus bachelor's/master's degree programs. Revisiting the current accreditation standards program may allow IPE to take a more prominent role in RT curricula. Copyright © 2017 by Daedalus Enterprises.

  4. [Correlation analysis of bone marrow edema degree and serum inflammatory factors change with knee joint pain symptoms in patients with bone contusion around the knee joint].

    PubMed

    Li, Songiun; An, Rongze; Wang, Zhaojie; Kuang, Lipeng; Tan, Weiyuan; Fang, Cunxun

    2014-05-01

    To explore the correlation between the degree of bone marrow edema (BME) and the content change of tumor necrosis factor alpha (TNF-alpha) and matrix metalloproteinase 3 (MMP-3) and the knee pain symptoms in patients with bone contusion around the knee joint. Thirty patients (30 knees) of bone contusion around the knee joint were chosen as the trial group between October 2009 and April 2012. According to visual analogue scale (VAS), 30 patients were divided into mild group (10 cases), moderate group (10 cases), and severe group (10 cases); according to MRI morphological changes, the patients were divided into type I group (12 cases), type II group (11 cases), and type III group (7 cases). Ten patients (10 knees) with soft tissue injury of the knee were chosen as control group. No significant difference was found (P > 0.05) in gender, age, causes, side, and admission time after injury between 2 groups. The serum contents of MMP-3 and TNF-alpha were detected and statistically analysed between different degrees of pain groups and between different degrees of BME groups. Correlation was analysed between BME and inflammatory factor changes and VAS score. The MMP-3 and TNF-alpha contents in trial group [(29.580 +/- 6.870) (microg/L and (23.750 +/- 7.096) ng/L] were significantly higher than those in control group [(8.219 +/- 1.355) microg/L and (6.485 +/- 1.168) ng/L] (t = 9.686, P = 0.000; t = 7.596, P =0.000). The MMP-3 and TNF-alpha contents in patients with different degrees of pain and BME were significantly higher than those in patients of control group (P < 0.05), and significant difference was found between patients with different degrees of pain (P < 0.05), but no significant difference between patients with different degrees of BME (P > 0.05). Multiple linear regression analysis showed that TNF-alpha content was significantly correlated with VAS score (P = 0.000). Knee pain symptoms are not related to the degree of BME in patients with bone contusion around the knee joint. Inflammatory factor TNF-alpha content is the main influence factor of knee joint pain symptoms.

  5. Application of the surface azimuthal electrical resistivity survey method to determine patterns of regional joint orientation in glacial tills

    USGS Publications Warehouse

    Carlson, D.

    2010-01-01

    Joints within unconsolidated material such as glacial till can be primary avenues for the flow of electrical charge, water, and contaminants. To facilitate the siting and design of remediation programs, a need exists to map anisotropic distribution of such pathways within glacial tills by determining the azimuth of the dominant joint set. The azimuthal survey method uses standard resistivity equipment with a Wenner array rotated about a fixed center point at selected degree intervals that yields an apparent resistivity ellipse. From this ellipse, joint set orientation can be determined. Azimuthal surveys were conducted at 21 sites in a 500-km2 (193 mi2) area around Milwaukee, Wisconsin, and more specifically, at sites having more than 30 m (98 ft) of glacial till (to minimize the influence of underlying bedrock joints). The 26 azimuthal surveys revealed a systematic pattern to the trend of the dominant joint set within the tills, which is approximately parallel to ice flow direction during till deposition. The average orientation of the joint set parallel with the ice flow direction is N77??E and N37??E for the Oak Creek and Ozaukee tills, respectively. The mean difference between average direct observation of joint set orientations and average azimuthal resistivity results is 8??, which is one fifth of the difference of ice flow direction between the Ozaukee and Oak Creek tills. The results of this study suggest that the surface azimuthal electrical resistivity survey method used for local in situ studies can be a useful noninvasive method for delineating joint sets within shallow geologic material for regional studies. Copyright ?? 2010 The American Association of Petroleum Geologists/Division of Environmental Geosciences. All rights reserved.

  6. Spacesuit mobility joints

    NASA Technical Reports Server (NTRS)

    Vykukal, H. C. (Inventor)

    1978-01-01

    Joints for use in interconnecting adjacent segments of an hermetically sealed spacesuit which have low torques, low leakage and a high degree of reliability are described. Each of the joints is a special purpose joint characterized by substantially constant volume and low torque characteristics. Linkages which restrain the joint from longitudinal distension and a flexible, substantially impermeable diaphragm of tubular configuration spanning the distance between pivotally supported annuli are featured. The diaphragms of selected joints include rolling convolutions for balancing the joints, while various joints include wedge-shaped sections which enhance the range of motion for the joints.

  7. Swivel Joint For Liquid Nitrogen

    NASA Technical Reports Server (NTRS)

    Milner, James F.

    1988-01-01

    Swivel joint allows liquid-nitrogen pipe to rotate through angle of 100 degree with respect to mating pipe. Functions without cracking hard foam insulation on lines. Pipe joint rotates on disks so mechanical stress not transmitted to thick insulation on pipes. Inner disks ride on fixed outer disks. Disks help to seal pressurized liquid nitrogen flowing through joint.

  8. Three-dimensional in vivo kinematics of the subtalar joint during dorsi-plantarflexion and inversion-eversion.

    PubMed

    Goto, Akira; Moritomo, Hisao; Itohara, Tomonobu; Watanabe, Tetsu; Sugamoto, Kazuomi

    2009-05-01

    It is difficult to determine the kinematics of the subtalar joint because of its anatomical and functional complexity. The purpose of the study was to clarify the 3D kinematics of the subtalar joint in vivo. Subjects were four healthy female volunteers. Magnetic resonance imaging (MRI) sequences were acquired in seven positions during dorsi-plantarflexion (DPF) and in 10 positions during inversion-eversion (IE) at intervals of 10 degrees. MRI data of the talus and calcaneus in the neutral position were superimposed on images of the other positions using voxel-based registration, and relative motions and axes of rotation were visualized and quantitatively calculated. The calcaneus always rotated from dorsolateral to medioplantar during DPF and IE, and the motion plane was very similar to that of the entire foot in IE. The axes of rotation of the calcaneus relative to the talus during DPF and IE had a very close spatial relationship, running obliquely from antero-dorso-medial to postero-planto-lateral and penetrating the talar neck. The rotation angle around each of these calcaneal axes tended to be greater in IE (20 degrees +/- 2 degrees) than in DPF (16 degrees +/- 3 degrees). In DPF, motion of the calcaneus relative to the talus occurred predominantly around maximum dorsiflexion and plantarflexion, with little movement observed at intermediate positions. During IE, the calcaneus exhibited uninterrupted motion related to foot movement. The subtalar joint is essentially a uniaxial joint with a motion plane almost identical to that of IE of the entire foot. Knowledge of normal subtalar kinematics may be helpful when evaluating pathologic conditions.

  9. Quantitative metal magnetic memory reliability modeling for welded joints

    NASA Astrophysics Data System (ADS)

    Xing, Haiyan; Dang, Yongbin; Wang, Ben; Leng, Jiancheng

    2016-03-01

    Metal magnetic memory(MMM) testing has been widely used to detect welded joints. However, load levels, environmental magnetic field, and measurement noises make the MMM data dispersive and bring difficulty to quantitative evaluation. In order to promote the development of quantitative MMM reliability assessment, a new MMM model is presented for welded joints. Steel Q235 welded specimens are tested along the longitudinal and horizontal lines by TSC-2M-8 instrument in the tensile fatigue experiments. The X-ray testing is carried out synchronously to verify the MMM results. It is found that MMM testing can detect the hidden crack earlier than X-ray testing. Moreover, the MMM gradient vector sum K vs is sensitive to the damage degree, especially at early and hidden damage stages. Considering the dispersion of MMM data, the K vs statistical law is investigated, which shows that K vs obeys Gaussian distribution. So K vs is the suitable MMM parameter to establish reliability model of welded joints. At last, the original quantitative MMM reliability model is first presented based on the improved stress strength interference theory. It is shown that the reliability degree R gradually decreases with the decreasing of the residual life ratio T, and the maximal error between prediction reliability degree R 1 and verification reliability degree R 2 is 9.15%. This presented method provides a novel tool of reliability testing and evaluating in practical engineering for welded joints.

  10. Phase 1 Program Joint Report

    NASA Technical Reports Server (NTRS)

    Nield, George C. (Editor); Vorobiev, Pavel Mikhailovich (Editor)

    1999-01-01

    This report consists of inputs from each of the Phase I Program Joint Working Groups. The Working Groups were tasked to describe the organizational structure and work processes that they used during the program, joint accomplishments, lessons learned, and applications to the International Space Station Program. This report is a top-level joint reference document that contains information of interest to both countries.

  11. Improved Automatically Locking/Unlocking Orthotic Knee Joint

    NASA Technical Reports Server (NTRS)

    Weddendorf, Bruce

    1995-01-01

    Proposed orthotic knee joint improved version of one described in "Automatically Locking/Unlocking Orthotic Knee Joint" (MFS-28633). Locks automatically upon initial application of radial force (wearer's weight) and unlocks automatically, but only when all loads (radial force and bending) relieved. Joints lock whenever wearer applies weight to knee at any joint angle between full extension and 45 degree bend. Both devices offer increased safety and convenience relative to conventional orthotic knee joints.

  12. Double degree master program: Optical Design

    NASA Astrophysics Data System (ADS)

    Bakholdin, Alexey; Kujawinska, Malgorzata; Livshits, Irina; Styk, Adam; Voznesenskaya, Anna; Ezhova, Kseniia; Ermolayeva, Elena; Ivanova, Tatiana; Romanova, Galina; Tolstoba, Nadezhda

    2015-10-01

    Modern tendencies of higher education require development of master programs providing achievement of learning outcomes corresponding to quickly variable job market needs. ITMO University represented by Applied and Computer Optics Department and Optical Design and Testing Laboratory jointly with Warsaw University of Technology represented by the Institute of Micromechanics and Photonics at The Faculty of Mechatronics have developed a novel international master double-degree program "Optical Design" accumulating the expertise of both universities including experienced teaching staff, educational technologies, and experimental resources. The program presents studies targeting research and professional activities in high-tech fields connected with optical and optoelectronics devices, optical engineering, numerical methods and computer technologies. This master program deals with the design of optical systems of various types, assemblies and layouts using computer modeling means; investigation of light distribution phenomena; image modeling and formation; development of optical methods for image analysis and optical metrology including optical testing, materials characterization, NDT and industrial control and monitoring. The goal of this program is training a graduate capable to solve a wide range of research and engineering tasks in optical design and metrology leading to modern manufacturing and innovation. Variability of the program structure provides its flexibility and adoption according to current job market demands and personal learning paths for each student. In addition considerable proportion of internship and research expands practical skills. Some special features of the "Optical Design" program which implements the best practices of both Universities, the challenges and lessons learnt during its realization are presented in the paper.

  13. Preliminary soft-tissue distraction versus checkrein ligament release after fasciectomy in the treatment of dupuytren proximal interphalangeal joint contractures.

    PubMed

    Craft, Randall O; Smith, Anthony A; Coakley, Brandon; Casey, William J; Rebecca, Alanna M; Duncan, Scott F M

    2011-11-01

    Checkrein ligament release for treatment of proximal interphalangeal joint Dupuytren contractures does not address the shortened arteries or deficient skin. The Digit Widget uses soft-tissue distraction to overcome these issues. This study compares checkrein ligament release after fasciectomy versus preliminary soft-tissue distraction, followed by operative release, for treatment of proximal interphalangeal joint Dupuytren contractures. The authors compared operative and postoperative characteristics of patients treated with either fasciectomy plus checkrein ligament release or Digit Widget distraction between 2001 and 2008. Seventeen patients (20 digits) underwent ligament release (mean contracture, 55.9 degrees); six of these 20 were reoperations. Thirteen patients (17 digits) underwent distraction (mean contracture, 67.6 degrees); 10 of 17 were reoperations. The 20 digits treated with fasciectomy plus ligament release had an average extension improvement of 31.4 degrees (range, -4 to 70 degrees). Digits treated with distraction had an average extension improvement of 53.4 degrees (range, 30 to 75 degrees) (p<0.001 versus ligament release). Three digits treated with distraction improved to full proximal interphalangeal extension. Initial contractures of 60 degrees or less treated by ligament release (n=12) or distraction (n=7) improved by means of 28.8 degrees and 47.7 degrees, respectively (p=0.048). Contractures greater than 60 degrees treated by ligament release (n=8) or distraction (n=10) improved by means of 35.3 degrees and 57.3 degrees, respectively (p=0.02). Soft-tissue distraction followed by operative release showed greater correction than Dupuytren fasciectomy plus checkrein ligament release. Therapeutic, III.

  14. Factors determining the level and changes in intra-articular pressure in the knee joint of the dog.

    PubMed Central

    Nade, S; Newbold, P J

    1983-01-01

    Intra-articular pressure levels were determined for joint positions throughout the normal physiological range of movement of dogs' knee joints. Change in joint position resulted in change in intra-articular pressure. It was demonstrated that intra-articular pressure is highest with the joint in the fully flexed position. Minimum pressure was recorded at a position between 80 degrees and 120 degrees. Minimum pressures were usually subatmospheric. The rate of change of joint position affected intra-articular pressure. The relationship of intra-articular pressure and joint position before and after full flexion demonstrated a hysteresis effect; the pressures were lower than for the same joint position before flexion. Maintenance of the joint in the fully flexed position for increasing periods of time between repeated movement cycles resulted in a similar reduction, of constant magnitude, in pressure between joint positions before and after each period of flexion. However, there was also a progressive decrease in pressure for all joint angles over the total number of movement cycles. There is a contribution to intra-articular pressure of joint capsular compliance and fluid movement into and out of the joint (both of which are time-dependent). The recording of intra-articular pressure in conscious, upright dogs revealed similar pressure levels to those measured in anaesthetized supine dogs. The major determinants of intra-articular pressure in normal dog knee joints include joint size, synovial fluid volume, position of joint, peri-articular tissue and joint anatomy, membrane permeability, capsular compliance, and movement of fluid into and out of the joint. Images Fig. 1 PMID:6875957

  15. Scaling NASA Applications to 1024 CPUs on Origin 3K

    NASA Technical Reports Server (NTRS)

    Taft, Jim

    2002-01-01

    The long and highly successful joint SGI-NASA research effort in ever larger SSI systems was to a large degree the result of the successful development of the MLP scalable parallel programming paradigm developed at ARC: 1) MLP scaling in real production codes justified ever larger systems at NAS; 2) MLP scaling on 256p Origin 2000 gave SGl impetus to productize 256p; 3) MLP scaling on 512 gave SGI courage to build 1024p O3K; and 4) History of MLP success resulted in IBM Star Cluster based MLP effort.

  16. Programming methodology for a general purpose automation controller

    NASA Technical Reports Server (NTRS)

    Sturzenbecker, M. C.; Korein, J. U.; Taylor, R. H.

    1987-01-01

    The General Purpose Automation Controller is a multi-processor architecture for automation programming. A methodology has been developed whose aim is to simplify the task of programming distributed real-time systems for users in research or manufacturing. Programs are built by configuring function blocks (low-level computations) into processes using data flow principles. These processes are activated through the verb mechanism. Verbs are divided into two classes: those which support devices, such as robot joint servos, and those which perform actions on devices, such as motion control. This programming methodology was developed in order to achieve the following goals: (1) specifications for real-time programs which are to a high degree independent of hardware considerations such as processor, bus, and interconnect technology; (2) a component approach to software, so that software required to support new devices and technologies can be integrated by reconfiguring existing building blocks; (3) resistance to error and ease of debugging; and (4) a powerful command language interface.

  17. Measurement of the passive stiffness of ankle joint in 3 DOF using stewart platform type ankle foot device.

    PubMed

    Nomura, Kenta; Yonezawa, Teru; Mizoguchi, Hiroshi; Takemura, Hiroshi

    2016-08-01

    This paper presents a method to measure the passive stiffness of an ankle joint in three degrees of freedom (DOF) under two motion speeds (1 Hz and 5 degree/s) using a developed Stewart platform-type device. The developed device can reproduce input motions of the foot in 6 DOF by controlling six pneumatic linear motion actuators. We used the device to measure the passive stiffness of an ankle joint undergoing three kinds of motion, namely dorsi-plantar flexion, inversion-eversion, and adduction-abduction. The measured values of the passive stiffness of the ankle joint in dorsiflexion that we obtained agreed well with that obtained in a previous study, indicating that the developed device is useful for measuring the passive stiffness of ankle joint. In addition, the developed device can be used to measure the stiffness in inversion-eversion and adduction-abduction motions as well, parameters that have never been measured. The results we obtained demonstrated certain interesting features as we varied both the direction and pace of motion (e.g., there were significant differences in the stiffness not only between adduction and abduction during the faster pace, but also between these and the other motions).

  18. Research Center for Optical Physics: Education and Technology for the 21st Century

    NASA Technical Reports Server (NTRS)

    2003-01-01

    During the past eleven years since its inception, RCOP has excelled in its two primary goals: 1) training of the scientists and engineers needed for the twenty-first century with special emphasis on underrepresented citizens and 2) research and technological development in areas of relevance to NASA. In the category of research training, as of May 2003, RCOP produced 36 Bachelors degrees, 25 Masters degrees, and 13 Doctoral degrees. Of these, all 36 Bachelors degrees, 16 of the Masters degrees and 9 of the Doctoral degrees were awarded to African Americans. Four of the Doctoral graduates and one of the Masters graduates are working at NASA Field Centers. RCOP has also provided research experiences to 130 undergraduate students and 22 high school students through a number of outreach programs held during the summer and the academic year. RCOP has also been crucial to the development of the Ph.D. program in physics at Hampton University by providing high quality research training and technical electives required for a Doctoral degree in physics. RCOP has also excelled in research and technological development. Since 1992, RCOP researchers have leveraged over 8 million dollars in additional research funding, published 152 papers in refereed journals and proceedings, and given 125 presentations at refereed international conferences in the United States and eight other countries. RCOP also developed numerous collaborations with other research centers, universities and industries. In recognition of this outstanding work, RCOP is the first research center in the United States invited to join the Joint Open Laboratory for Laser Crystals and Precise Laser Systems headed by Dr. Alexander Kaminiskii of the Russian Academy of Sciences.

  19. The CELSS Antarctic Analog Project: an advanced life support testbed at the Amundsen-Scott South Pole Station, Antarctica.

    PubMed

    Straight, C L; Bubenheim, D L; Bates, M E; Flynn, M T

    1994-01-01

    The Controlled Ecological Life Support System (CELSS) Antarctic Analog Project (CAAP) is a joint endeavor between the National Science Foundation, Office of Polar Programs (NSF-OPP) and the National Aeronautics and Space Administration (NASA). Its fundamental objective is to develop, deploy and operate a testbed of NASA CELSS technologies and life support approaches at the Amundsen-Scott South Pole Station, located at latitude 90 degrees S, longitude 0 degrees. The goal of NASA's CELSS Program is to develop technologies and systems that will allow spacefaring scientists and explorers to carry out long duration extraterrestrial missions, leading ultimately to permanent habitation of the Solar System, without total dependence on a costly resupply system. A CELSS would do this by providing regenerated life support materials (air, food and water) and by processing "waste" materials into useful resources. This will be accomplished using biological and physical/chemical techniques in a nearly closed environmental habitation system. CELSS technologies also have great implications for application to terrestrial systems with intrinsic transferability to society at large. The CELSS Program intends to provide opportunities for the transfer of these systems and technologies outside the US Space Program, to applications within the American economy as space technology spin-offs.

  20. Adaptive velocity-based six degree of freedom load control for real-time unconstrained biomechanical testing.

    PubMed

    Lawless, I M; Ding, B; Cazzolato, B S; Costi, J J

    2014-09-22

    Robotic biomechanics is a powerful tool for further developing our understanding of biological joints, tissues and their repair. Both velocity-based and hybrid force control methods have been applied to biomechanics but the complex and non-linear properties of joints have limited these to slow or stepwise loading, which may not capture the real-time behaviour of joints. This paper presents a novel force control scheme combining stiffness and velocity based methods aimed at achieving six degree of freedom unconstrained force control at physiological loading rates. Copyright © 2014 Elsevier Ltd. All rights reserved.

  1. Joint Space Operations Center (JSpOC) Mission System Increment 3 (JMS Inc 3)

    DTIC Science & Technology

    2016-03-01

    2016 Major Automated Information System Annual Report Joint Space Operations Center (JSpOC) Mission System Increment 3 (JMS Inc 3) Defense...1725 DSN Phone: DSN Fax: Date Assigned: May 16, 2014 Program Information Program Name Joint Space Operations Center (JSpOC) Mission System...approved program baseline; therefore, no Original Estimate has been established. JMS Inc 3 2016 MAR UNCLASSIFIED 4 Program Description The Joint Space

  2. Tissue Engineering Initiative

    DTIC Science & Technology

    2002-08-01

    evaluate functionality, the FDP/SIS and FDS were independently pulled to determine the degree of distal interphalangeal (DIP) joint motion contributed by...each. In three digits the distal phalanx moved similarly whether pulling on the FDP/SIS or the FDS tendon. This suggests some scarring/adhesions between... pulled to determine the degree of distal interphalangeal (DIP) joint motion contributed by each. In three digits the distal phalanx moved similarly

  3. Spacesuit mobility knee joints

    NASA Technical Reports Server (NTRS)

    Vykukal, H. C. (Inventor)

    1979-01-01

    Pressure suit mobility joints are for use in interconnecting adjacent segments of an hermetically sealed spacesuit in which low torques, low leakage and a high degree of reliability are required. Each of the joints is a special purpose joint characterized by substantially constant volume and low torque characteristics and includes linkages which restrain the joint from longitudinal distension and includes a flexible, substantially impermeable diaphragm of tubular configuration spanning the distance between pivotally supported annuli. The diaphragms of selected joints include rolling convolutions for balancing the joints, while various joints include wedge-shaped sections which enhance the range of motion for the joints.

  4. Anatomical and biomechanical investigations of the iliotibial tract.

    PubMed

    Birnbaum, K; Siebert, C H; Pandorf, T; Schopphoff, E; Prescher, A; Niethard, F U

    2004-12-01

    Divergent descriptions of the anatomic location and biomechanical function of the iliotibial tract (IT) can be found in the literature. This study attempted to obtain exact data regarding the anatomic course and material characteristics including the biomechanical properties of this structure. The following were its aims: (1) anatomical investigations of the IT; (2) mechanical properties of the IT; (3) femoral head centralizing force of the IT and subligamentous forces in the height of the greater trochanter in different joint positions by using a custom-made measuring prosthesis and a subligamentous positioned sensor; (4) construction of a finite element model of the proximal femur including the IT and measuring the femoral neck angle under variation. The hip joints and IT in a total of 18 unfixed corpses were evaluated. We studied the anatomic relationship to surrounding structures, as well as the material properties with the help of tensile strength testing utilizing an uniaxial apparatus. During the test, a load-displacement curve was registered, documenting the maximum load and deformation of the IT. To measure the subligamentous pressure at the height of the greater trochanter, a custom-made sensor with a power-recording instrument was constructed. Furthermore, an altered hip prosthesis with a pressure gauge at the height of the femoral neck was used to measure the forces which are directed at the acetabulum. The investigations were done in neutral-0 position and ab/adduction of the hip joint of the unfixed corpse. In addition, we varied the femoral neck angle between 115 degrees and 155 degrees in 5 degrees steps. To confirm the subligamentous forces, we did the same measurements intraoperatively at the height of the greater trochanter before and after hip joint replacement in 12 patients. We constructed a finite element model of the proximal femur and considering the IT. The acquisition of the data was done at physiological (128 degrees), varus (115 degrees), and valgus (155 degrees) femoral neck angles. The influencing forces of the IT at the height of the greater trochanter and the forces at the femoral head or the acetabulum could be measured. Our anatomical investigations revealed a splitting of the IT into a superficial and a deep portion, which covers the tensor fasciae latae. The tensor fasciae latae has an insertion on the IT. The IT continues down the femur, passing over the greater trochanter without developing an actual fixation to the bone. Part of the insertion of the gluteus maximus radiates into the IT. The IT passes over the vastus lateralis and inserts at the infracondylar tubercle of the tibia or Gerdy's tubercle, at the head of the fibula, as well as at the lateral intermuscular septum. Portions also insert on the transverse and longitudinal retinaculum of the patella. Concerning the material properties of the IT, we found a structural stiffness of 17 N/mm extension on average (D = 17 N/mm). The subligamentous measurements at the height of the greater trochanter in the unfixed corpse and intraoperatively during hip joint replacement showed an increase of the forces during adduction and a decrease during abduction of the hip joint. We found thereby a maximum increase up to 106 N with 40 degrees adduction. Concerning the femoral neck angle, we can state that valgus leads to lower subligamentous forces and varus to higher subligamentous forces. The forces directed at the acetabulum, which were measured by the prosthesis with a sensor along the femoral neck, showed a decrease with varus angles and an increase with valgus angles. The highest force of 624 N was measured with 40 degrees adduction and an angle of 155 degrees. The finite element model of the proximal femur showed a sole hip joint-centralizing force of the IT of 655 N with a femoral neck angle of 128 degrees after subtraction of the gluteal muscle force and the body weight. At 115 degrees, we found an increase up to 997 N and a decrease to 438 N at 155 degrees. Concerning the resulting forces in the acetabulum, we found opposite forces in comparison with the force of the IT at the height of the greater trochanter: at 115 degrees, a femoral head-centralizing force of 1601 N; at 128 degrees, 2360 N; and at 155 degrees, 2422 N. By our investigations, we can approximately prove the hip joint-centralizing force of the IT. By variation of the femoral neck angle and the position of the hip joint, we can predict the subligamentous force of the IT and the resulting force at the femoral head or at the acetabulum. The intraoperative measurement of the subligamentous forces of the IT is a good monitoring mechanism for the persistent hip-centralizing function of the IT in the course of hip joint replacement. The surgeon has the opportunity to check the stability of the hip joint after replacement. The finite element model gives the opportunity to check the divergent relative strength by variation of the femoral neck angle and the tension of the IT. In this way, the changes in the forces induced by a displacement osteotomy could be estimated preoperatively.

  5. Joint Exercise Program: DOD Needs to Take Steps to Improve the Quality of Funding Data

    DTIC Science & Technology

    2017-02-01

    technology systems—the Joint Training Information Management System (JTIMS) and the Execution Management System—to manage the execution of the Joint...Exercise Program, but does not have assurance that funding execution data in the Execution Management System are reliable. JTIMS is the system of record...for the Joint Exercise Program that combatant commanders use to plan and manage their joint training exercises. GAO observed significant variation

  6. The influence of muscle length on the fatigue-related reduction in joint range of motion of the human dorsiflexors.

    PubMed

    Cheng, Arthur J; Davidson, Andrew W; Rice, Charles L

    2010-06-01

    The fatigue-related reduction in joint range of motion (ROM) during dynamic contraction tasks may be related to muscle length-dependent alterations in torque and contractile kinetics, but this has not been systematically explored previously. Twelve young men performed a repetitive voluntary muscle shortening contraction task of the dorsiflexors at a contraction load of 30% of maximum voluntary isometric contraction (MVC) torque, until total 40 degrees ROM had decreased by 50% at task failure (POST) to 20 degrees ROM. At both a short (5 degrees dorsiflexion) and long muscle length (35 degrees plantar flexion joint angle relative to a 0 degrees neutral ankle joint position), voluntary activation, MVC torque, and evoked tibialis anterior contractile properties of a 52.8 Hz high-frequency isometric tetanus [peak evoked torque, maximum rate of torque development (MRTD), maximum rate of relaxation (MRR)] were evaluated at baseline (PRE), at POST, and up to 10 min of recovery. At POST, we measured similar fatigue-related reductions in torque (voluntary and evoked) and slowing of contractile kinetics (MRTD and MRR) at both the short and long muscle lengths. Thus, the fatigue-related reduction in ROM could not be explained by length-dependent fatigue. Although torque (voluntary and evoked) at both muscle lengths was depressed and remained blunted throughout the recovery period, this was not related to the rapid recovery of ROM at 0.5 min after task failure. The reduction in ROM, however, was strongly related to the reduction in joint angular velocity (R(2) = 0.80) during the fatiguing task, although additional factors cannot yet be overlooked.

  7. Three-dimensional Reconstruction of Scar Contracture-bearing Axilla and Digital Webs Using the Square Flap Method.

    PubMed

    Huang, Chenyu; Ogawa, Rei

    2014-05-01

    Joint scar contractures are characterized by tight bands of soft tissue that bridge the 2 ends of the joint like a web. Classical treatment methods such as Z-plasties are mainly based on 2-dimensional designs. Our square flap method is an alternative surgical method that restores the span of the web in a stereometric fashion, thereby reconstructing joint function. In total, 20 Japanese patients with joint scar contractures on the axillary (n = 10) or first digital web (n = 10) underwent square flap surgery. The maximum range of motion and commissure length were measured before and after surgery. A theoretical stereometric geometrical model of the square flap was established to compare it to the classical single (60 degree), 4-flap (45 degree), and 5-flap (60 degree) Z-plasties in terms of theoretical web reconstruction efficacy. All cases achieved 100% contracture release. The maximum range of motion and web space improved after square flap surgery (P = 0.001). Stereometric geometrical modeling revealed that the standard square flap (α = 45 degree; β = 90 degree) yields a larger flap area, length/width ratio, and postsurgical commissure length than the Z-plasties. It can also be adapted by varying angles α and β, although certain angle thresholds must be met to obtain the stereometric advantages of this method. When used to treat joint scar contractures, the square flap method can fully span the web space in a stereometric manner, thus yielding a close-to-original shape and function. Compared with the classical Z-plasties, it also provides sufficient anatomical blood supply while imposing the least physiological tension on the adjacent skin.

  8. Three-dimensional Reconstruction of Scar Contracture-bearing Axilla and Digital Webs Using the Square Flap Method

    PubMed Central

    Huang, Chenyu

    2014-01-01

    Background: Joint scar contractures are characterized by tight bands of soft tissue that bridge the 2 ends of the joint like a web. Classical treatment methods such as Z-plasties are mainly based on 2-dimensional designs. Our square flap method is an alternative surgical method that restores the span of the web in a stereometric fashion, thereby reconstructing joint function. Methods: In total, 20 Japanese patients with joint scar contractures on the axillary (n = 10) or first digital web (n = 10) underwent square flap surgery. The maximum range of motion and commissure length were measured before and after surgery. A theoretical stereometric geometrical model of the square flap was established to compare it to the classical single (60 degree), 4-flap (45 degree), and 5-flap (60 degree) Z-plasties in terms of theoretical web reconstruction efficacy. Results: All cases achieved 100% contracture release. The maximum range of motion and web space improved after square flap surgery (P = 0.001). Stereometric geometrical modeling revealed that the standard square flap (α = 45 degree; β = 90 degree) yields a larger flap area, length/width ratio, and postsurgical commissure length than the Z-plasties. It can also be adapted by varying angles α and β, although certain angle thresholds must be met to obtain the stereometric advantages of this method. Conclusions: When used to treat joint scar contractures, the square flap method can fully span the web space in a stereometric manner, thus yielding a close-to-original shape and function. Compared with the classical Z-plasties, it also provides sufficient anatomical blood supply while imposing the least physiological tension on the adjacent skin. PMID:25289342

  9. Ankle joint pressure changes in a pes cavovarus model: supramalleolar valgus osteotomy versus lateralizing calcaneal osteotomy.

    PubMed

    Schmid, Timo; Zurbriggen, Sebastian; Zderic, Ivan; Gueorguiev, Boyko; Weber, Martin; Krause, Fabian G

    2013-09-01

    A fixed cavovarus foot deformity can be associated with anteromedial ankle arthrosis due to elevated medial joint contact stresses. Supramalleolar valgus osteotomies (SMOT) and lateralizing calcaneal osteotomies (LCOT) are commonly used to treat symptoms by redistributing joint contact forces. In a cavovarus model, the effects of SMOT and LCOT on the lateralization of the center of force (COF) and reduction of the peak pressure in the ankle joint were compared. A previously published cavovarus model with fixed hindfoot varus was simulated in 10 cadaver specimens. Closing wedge supramalleolar valgus osteotomies 3 cm above the ankle joint level (6 and 11 degrees) and lateral sliding calcaneal osteotomies (5 and 10 mm displacement) were analyzed at 300 N axial static load (half body weight). The COF migration and peak pressure decrease in the ankle were recorded using high-resolution TekScan pressure sensors. A significant lateral COF shift was observed for each osteotomy: 2.1 mm for the 6 degrees (P = .014) and 2.3 mm for the 11 degrees SMOT (P = .010). The 5 mm LCOT led to a lateral shift of 2.0 mm (P = .042) and the 10 mm LCOT to a shift of 3.0 mm (P = .006). Comparing the different osteotomies among themselves no significant differences were recorded. No significant anteroposterior COF shift was seen. A significant peak pressure reduction was recorded for each osteotomy: The SMOT led to a reduction of 29% (P = .033) for the 6 degrees and 47% (P = .003) for the 11 degrees osteotomy, and the LCOT to a reduction of 41% (P = .003) for the 5 mm and 49% (P = .002) for the 10 mm osteotomy. Similar to the COF lateralization no significant differences between the osteotomies were seen. LCOT and SMOT significantly reduced anteromedial ankle joint contact stresses in this cavovarus model. The unloading effects of both osteotomies were equivalent. More correction did not lead to significantly more lateralization of the COF or more reduction of peak pressure but a trend was seen. In patients with fixed cavovarus feet, both SMOT and LCOT provided equally good redistribution of elevated ankle joint contact forces. Increasing the amount of displacement did not seem to equally improve the joint pressures. The site of osteotomy could therefore be chosen on the basis of surgeon's preference, simplicity, or local factors in case of more complex reconstructions.

  10. Effect of Preactivation on Torque Enhancement by the Stretch-Shortening Cycle in Knee Extensors

    PubMed Central

    Fukutani, Atsuki; Misaki, Jun; Isaka, Tadao

    2016-01-01

    The stretch-shortening cycle is one of the most interesting topics in the field of sport sciences, because the performance of human movement is enhanced by the stretch-shortening cycle (eccentric contraction). The purpose of the present study was to examine whether the influence of preactivation on the torque enhancement by stretch-shortening cycle in knee extensors. Twelve men participated in this study. The following three conditions were conducted for knee extensors: (1) concentric contraction without preactivation (CON), (2) concentric contraction with eccentric preactivation (ECC), and (3) concentric contraction with isometric preactivation (ISO). Muscle contractions were evoked by electrical stimulation to discard the influence of neural activity. The range of motion of the knee joint was set from 80 to 140 degrees (full extension = 180 degrees). Angular velocities of the concentric and eccentric contractions were set at 180 and 90 degrees/s, respectively. In the concentric contraction phase, joint torques were recorded at 85, 95, and 105 degrees, and they were compared among the three conditions. In the early phase (85 degrees) of concentric contraction, the joint torque was larger in the ECC and ISO conditions than in the CON condition. However, these clear differences disappeared in the later phase (105 degrees) of concentric contraction. The results showed that joint torque was clearly different among the three conditions in the early phase whereas this difference disappeared in the later phase. Thus, preactivation, which is prominent in the early phase of contractions, plays an important role in torque enhancement by the stretch-shortening cycle in knee extensors. PMID:27414804

  11. BIOMECHANICS AND PATHOMECHANICS OF THE PATELLOFEMORAL JOINT

    PubMed Central

    2016-01-01

    The patellofemoral joint is a joint that can be an area of concern for athletes of various sports and ages. The joint is somewhat complex with multiple contact points and numerous tissues that attach to the patella. Joint forces are variable and depend on the degree of knee flexion and whether the foot is in contact with the ground. The sports medicine specialist must have a good working knowledge of the anatomy and biomechanics of the patellofemoral joint in order to treat it effectively. PMID:27904787

  12. Development of a sensor coordinated kinematic model for neural network controller training

    NASA Technical Reports Server (NTRS)

    Jorgensen, Charles C.

    1990-01-01

    A robotic benchmark problem useful for evaluating alternative neural network controllers is presented. Specifically, it derives two camera models and the kinematic equations of a multiple degree of freedom manipulator whose end effector is under observation. The mapping developed include forward and inverse translations from binocular images to 3-D target position and the inverse kinematics of mapping point positions into manipulator commands in joint space. Implementation is detailed for a three degree of freedom manipulator with one revolute joint at the base and two prismatic joints on the arms. The example is restricted to operate within a unit cube with arm links of 0.6 and 0.4 units respectively. The development is presented in the context of more complex simulations and a logical path for extension of the benchmark to higher degree of freedom manipulators is presented.

  13. The Emergence of Nonverbal Joint Attention and Requesting Skills in Young Children with Autism

    ERIC Educational Resources Information Center

    Paparella, Tanya; Goods, Kelly Stickles; Freeman, Stephanny; Kasari, Connie

    2011-01-01

    Joint attention (JA) skills are deficient in children with autism; however, children with autism seem to vary in the degree to which they display joint attention. Joint attention skills refer to verbal and nonverbal skills used to share experiences with others. They include gestures such as pointing, coordinated looks between objects and people,…

  14. Towards an Analytical Framework for Understanding the Development of a Quality Assurance System in an International Joint Programme

    ERIC Educational Resources Information Center

    Zheng, Gaoming; Cai, Yuzhuo; Ma, Shaozhuang

    2017-01-01

    This paper intends to construct an analytical framework for understanding quality assurance in international joint programmes and to test it in a case analysis of a European--Chinese joint doctoral degree programme. The development of a quality assurance system for an international joint programme is understood as an institutionalization process…

  15. Evaluating the contribution of a neural component of ankle joint resistive torque in patients with stroke using a manual device.

    PubMed

    Kobayashi, Toshiki; Leung, Aaron K L; Akazawa, Yasushi; Hutchins, Stephen W

    2011-01-01

    To investigate the methodology using a manual ankle joint resistive torque measurement device to evaluate the contribution of the neural component of ankle joint resistive torque in patients with stroke. Within-subject comparison to compare the ankle joint resistive torque between fast and slow stretching conditions. Ten patients with stroke participated in this study. The incremental ratio of ankle joint resistive torque at the ankle angular position of 5degrees dorsiflexion under the fast stretching condition in comparison to the slow one was calculated in each patient. A significant increase (p<0.01) in the ankle joint resistive torque was demonstrated under the fast stretching condition in comparison to the slow one in all patients and the mean ankle joint resistive torque was 4.6 (SD=1.7) Nm under the slow stretching condition, while it was 8.4 (SD=4.1) Nm under the fast stretching condition at the ankle angular position of 5 degrees dorsiflexion. The incremental ratio ranged from 9.4-139.3% among the patients. The results of this study demonstrated the potential advantage of the device to evaluate the contribution of the neural component of ankle joint resistive torque.

  16. 15 CFR 296.8 - Joint venture agreement.

    Code of Federal Regulations, 2011 CFR

    2011-01-01

    ... OF STANDARDS AND TECHNOLOGY, DEPARTMENT OF COMMERCE NIST EXTRAMURAL PROGRAMS TECHNOLOGY INNOVATION PROGRAM General § 296.8 Joint venture agreement. NIST shall not issue a TIP award to a joint venture and... joint venture agreement has been executed by all of the joint venture members and approved by NIST. ...

  17. 15 CFR 296.8 - Joint venture agreement.

    Code of Federal Regulations, 2012 CFR

    2012-01-01

    ... OF STANDARDS AND TECHNOLOGY, DEPARTMENT OF COMMERCE NIST EXTRAMURAL PROGRAMS TECHNOLOGY INNOVATION PROGRAM General § 296.8 Joint venture agreement. NIST shall not issue a TIP award to a joint venture and... joint venture agreement has been executed by all of the joint venture members and approved by NIST. ...

  18. 15 CFR 296.8 - Joint venture agreement.

    Code of Federal Regulations, 2013 CFR

    2013-01-01

    ... OF STANDARDS AND TECHNOLOGY, DEPARTMENT OF COMMERCE NIST EXTRAMURAL PROGRAMS TECHNOLOGY INNOVATION PROGRAM General § 296.8 Joint venture agreement. NIST shall not issue a TIP award to a joint venture and... joint venture agreement has been executed by all of the joint venture members and approved by NIST. ...

  19. 15 CFR 296.8 - Joint venture agreement.

    Code of Federal Regulations, 2014 CFR

    2014-01-01

    ... OF STANDARDS AND TECHNOLOGY, DEPARTMENT OF COMMERCE NIST EXTRAMURAL PROGRAMS TECHNOLOGY INNOVATION PROGRAM General § 296.8 Joint venture agreement. NIST shall not issue a TIP award to a joint venture and... joint venture agreement has been executed by all of the joint venture members and approved by NIST. ...

  20. 15 CFR 296.8 - Joint venture agreement.

    Code of Federal Regulations, 2010 CFR

    2010-01-01

    ... OF STANDARDS AND TECHNOLOGY, DEPARTMENT OF COMMERCE NIST EXTRAMURAL PROGRAMS TECHNOLOGY INNOVATION PROGRAM General § 296.8 Joint venture agreement. NIST shall not issue a TIP award to a joint venture and... joint venture agreement has been executed by all of the joint venture members and approved by NIST. ...

  1. In arthritis the Doppler based degree of hypervascularisation shows a positive correlation with synovial leukocyte count and distinguishes joints with leukocytes greater and less than 5/nL.

    PubMed

    Löffler, Christian; Sattler, Horst; Peters, Lena; Tuleweit, Anika; Löffler, Uta; Wadsack, Daniel; Uppenkamp, Michael; Bergner, Raoul

    2016-10-01

    Power Doppler ultrasound is used to assess joint vascularity in acute arthritis. PDUS signals have been correlated with synovial histology and bone deterioration. Little is known about the correlation between power Doppler signals and synovial white blood count. In our study, we analyzed power Doppler signals in inflammatory joint diseases including gout, calcium pyrophosphate deposition disease, rheumatoid arthritis, spondyloarthritis and others and correlated power Doppler signals with synovial white blood count and with serologic markers of inflammation. We retrospectively evaluated 194 patients with arthritis. All patients underwent joint sonography, power Doppler ultrasound, synovial fluid analysis and blood examination of C-reactive protein and erythrocyte sedimentation rate. Correlation analyses (Spearman and Pearson), Chi(2) test, t-tests, a unifactorial ANOVA and regression analyses were applied. Hypervascularisation in power Doppler was most prominent in gout and calcium pyrophosphate deposition disease. Spondyloarthritis and non-inflammatory joint diseases presented with low degrees of hypervascularisation. Mean synovial white blood count did not differ significantly between crystal-related arthritides, rheumatoid arthritis, spondyloarthritis or other inflammatory joint diseases. There was a positive but weak correlation between power Doppler signals and synovial white blood count (P<0.001, rs=0.283), erythrocyte sedimentation rate (P<0.001, rs=0.387) and C-reactive protein (P<0.001, rs=0.373) over all diagnoses. This was especially relevant in rheumatoid arthritis (P<0.01, rs=0.479). Power Doppler degrees 0 and 1 were able to predict synovial leukocytes<5/nL, degrees 2 and 3 predict leukocytes≥5/nL (P<0.001). Copyright © 2016 Société française de rhumatologie. Published by Elsevier SAS. All rights reserved.

  2. Composition and method for brazing graphite to graphite

    DOEpatents

    Taylor, Albert J.; Dykes, Norman L.

    1984-01-01

    The present invention is directed to a brazing material for joining graphite structures that can be used at temperatures up to about 2800.degree. C. The brazing material formed of a paste-like composition of hafnium carbide and uranium oxide with a thermosetting resin. The uranium oxide is converted to uranium dicarbide during the brazing operation and then the hafnium carbide and uranium dicarbide form a liquid phase at a temperature about 2600.degree. C. with the uranium diffusing and vaporizing from the joint area as the temperature is increased to about 2800.degree. C. so as to provide a brazed joint consisting essentially of hafnium carbide. This brazing temperature for hafnium carbide is considerably less than the eutectic temperature of hafnium carbide of about 3150.degree. C. The brazing composition also incorporates the thermosetting resin so that during the brazing operation the graphite structures may be temporarily bonded together by thermosetting the resin so that machining of the structures to final dimensions may be completed prior to the completion of the brazing operation. The resulting brazed joint is chemically and thermally compatible with the graphite structures joined thereby and also provides a joint of sufficient integrity so as to at least correspond with the strength and other properties of the graphite.

  3. Effects of joint attention on long-term memory in 9-month-old infants: an event-related potentials study.

    PubMed

    Kopp, Franziska; Lindenberger, Ulman

    2011-07-01

    Joint attention develops during the first year of life but little is known about its effects on long-term memory. We investigated whether joint attention modulates long-term memory in 9-month-old infants. Infants were familiarized with visually presented objects in either of two conditions that differed in the degree of joint attention (high versus low). EEG indicators in response to old and novel objects were probed directly after the familiarization phase (immediate recognition), and following a 1-week delay (delayed recognition). In immediate recognition, the amplitude of positive slow-wave activity was modulated by joint attention. In the delayed recognition, the amplitude of the Pb component differentiated between high and low joint attention. In addition, the positive slow-wave amplitude during immediate and delayed recognition correlated with the frequency of infants' looks to the experimenter during familiarization. Under both high- and low-joint-attention conditions, the processing of unfamiliar objects was associated with an enhanced Nc component. Our results show that the degree of joint attention modulates EEG during immediate and delayed recognition. We conclude that joint attention affects long-term memory processing in 9-month-old infants by enhancing the relevance of attended items. © 2010 Blackwell Publishing Ltd.

  4. Relationship Between the Number of Clinical Sites in Radiography Programs and Job Placement Rates of Graduates.

    PubMed

    Harrell, Angela; Matthews, Eric

    2016-07-01

    To determine whether a relationship exists between the number of clinical sites available in radiography programs accredited by the Joint Review Committee on Education in Radiologic Technology and the job placement rates of graduates. We performed a secondary analysis of data on job placement rates and the number of clinical sites available in 438 degree-granting radiography programs from January 2015 to March 2015. A weak, negative, nonsignificant correlation existed between the number of clinical sites and the job placement rate (Spearman's rho = -.113, n = 438, P = .018). The coefficient of determination was 1.28%.Discussion Research evaluating factors contributing to graduate employability is limited but indicates no need for radiography program administrators to adjust clinical site numbers solely on the basis of improving graduate employability. The number of clinical sites available in a radiography program is not related to the job placement rate of its graduates. ©2016 American Society of Radiologic Technologists.

  5. Effect of proprioception training on knee joint position sense in female team handball players.

    PubMed

    Pánics, G; Tállay, A; Pavlik, A; Berkes, I

    2008-06-01

    A number of studies have shown that proprioception training can reduce the risk of injuries in pivoting sports, but the mechanism is not clearly understood. To determine the contributing effects of propioception on knee joint position sense among team handball players. Prospective cohort study. Two professional female handball teams were followed prospectively for the 2005-6 season. 20 players in the intervention team followed a prescribed proprioceptive training programme while 19 players in the control team did not have a specific propioceptive training programme. The coaches recorded all exposures of the individual players. The location and nature of injuries were recorded. Joint position sense (JPS) was measured by a goniometer on both knees in three angle intervals, testing each angle five times. Assessments were performed before and after the season by the same examiner for both teams. In the intervention team a third assessment was also performed during the season. Complete data were obtained for 15 subjects in the intervention team and 16 in the control team. Absolute error score, error of variation score and SEM were calculated and the results of the intervention and control teams were compared. The proprioception sensory function of the players in the intervention team was significantly improved between the assessments made at the start and the end of the season (mean (SD) absolute error 9.78-8.21 degrees (7.19-6.08 degrees ) vs 3.61-4.04 degrees (3.71-3.20 degrees ), p<0.05). No improvement was seen in the sensory function in the control team between the start and the end of the season (mean (SD) absolute error 6.31-6.22 degrees (6.12-3.59 degrees ) vs 6.13-6.69 degrees (7.46-6.49 degrees ), p>0.05). This is the first study to show that proprioception training improves the joint position sense in elite female handball players. This may explain the effect of neuromuscular training in reducing the injury rate.

  6. Effects of Joint Attention on Long-Term Memory in 9-Month-Old Infants: An Event-Related Potentials Study

    ERIC Educational Resources Information Center

    Kopp, Franziska; Lindenberger, Ulman

    2011-01-01

    Joint attention develops during the first year of life but little is known about its effects on long-term memory. We investigated whether joint attention modulates long-term memory in 9-month-old infants. Infants were familiarized with visually presented objects in either of two conditions that differed in the degree of joint attention (high…

  7. Brief Report: High and Low Level Initiations of Joint Attention, and Response to Joint Attention--Differential Relationships with Language and Imitation

    ERIC Educational Resources Information Center

    Pickard, Katherine E.; Ingersoll, Brooke R.

    2015-01-01

    Frequency of high-level (showing/pointing) and low-level (coordinated gaze shifts) behaviors on the Early Social Communication Scales are often used as a measure of joint attention initiations (IJA). This study examined the degree to which these skills and response to joint attention (RJA; e.g. gaze following) were differentially related to…

  8. Experimental Studies on the Mechanical Behaviour of Rock Joints with Various Openings

    NASA Astrophysics Data System (ADS)

    Li, Y.; Oh, J.; Mitra, R.; Hebblewhite, B.

    2016-03-01

    The mechanical behaviour of rough joints is markedly affected by the degree of joint opening. A systematic experimental study was conducted to investigate the effect of the initial opening on both normal and shear deformations of rock joints. Two types of joints with triangular asperities were produced in the laboratory and subjected to compression tests and direct shear tests with different initial opening values. The results showed that opened rock joints allow much greater normal closure and result in much lower normal stiffness. A semi-logarithmic law incorporating the degree of interlocking is proposed to describe the normal deformation of opened rock joints. The proposed equation agrees well with the experimental results. Additionally, the results of direct shear tests demonstrated that shear strength and dilation are reduced because of reduced involvement of and increased damage to asperities in the process of shearing. The results indicate that constitutive models of rock joints that consider the true asperity contact area can be used to predict shear resistance along opened rock joints. Because rock masses are loosened and rock joints become open after excavation, the model suggested in this study can be incorporated into numerical procedures such as finite-element or discrete-element methods. Use of the model could then increase the accuracy and reliability of stability predictions for rock masses under excavation.

  9. Experimental cryo-irrigation of the knee joint.

    PubMed

    Chen, S C; Helal, B; Revell, P A; Brocklehurst, R; Currey, H L

    1986-10-01

    Experiments have been carried out to test the feasibility of using cryo-irrigation as a means of ablating the synovium in the rheumatoid knee joint. Cryo-irrigation was performed by a cooling machine and pump, which circulated cold 200/10 centistoke (cSt) silicone through the knee joint of rabbits anaesthetised with intravenous (IV) 'Saffan'. Fluid left the joint at -5 to -10 degrees C. Sixteen normal New Zealand rabbits received cryo-irrigation of one knee joint for 10-20 minutes and were killed at one day, and one, two, and 12 weeks thereafter. Judged by radioactive sulphate incorporation there was no impairment of chondrocyte function in the articular cartilage of irrigated joints. Histological examination showed mild synovitis and some loss of staining of superficial cartilage in 6/16 irrigated joints (v 1/16 control joints). Similar treatment of rabbit joints in which the Glynn model of synovitis had been induced showed marked reduction of synovitis 14-45 days after silicone treatment. Nine of 26 animals in which synovitis was induced in both knees and cryo-irrigation performed in one knee died either immediately postoperatively or during the next week. These deaths remain unexplained. A single dog received cryo-irrigation of one knee (-6 to -9 degrees C for 22 min) and remained perfectly well up to sacrifice at six months, when the joint appeared histologically completely normal.

  10. Shop test of the 501F; A 150 MW combustion turbine

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Entenmann, D.T.; North, W.E.; Fukue, I.

    1991-10-01

    The 501F is a 150 MW-class 60 Hz engine jointly developed by Westinghouse Electric Corporation and Mitsubishi Heavy Industries, Ltd. This paper describes the full-load shop test program for the prototype engine, as carried out in Takasago, Japan. The shop test included a full range of operating conditions, from startup through full load at the 1260{degrees} C (2300{degrees} F) design turbine inlet temperature. The engine was prepared with more than 1500 instrumentation points to monitor flow path characteristics, metal temperatures, displacements, pressures, cooling circuit characteristics, strains, sound pressure levels, and exhaust emissions. The results of this shop test indicate themore » new 501F engine design and development effort to be highly successful. The engine exceeds power and overall efficiency expectations, thus verifying the new concepts and design improvements.« less

  11. The Relationship Between Osteoarthritis of the Lumbar Facet Joints and Lumbosacropelvic Morphology.

    PubMed

    Sahin, Mehmet Sukru; Ergün, Adviye; Aslan, Akın

    2015-10-01

    Cross-sectional study. To investigate the relation between lumbosacropelvic morphology and the presence and degree of facet joint degeneration. Osteoarthritis of the facet joints is one of the most common degenerative changes in the spine. It is considered to be formed secondary to repetitive stress or trauma and spinal deformity with secondary overload. The cause(s) of facet joints osteoarthritis, however, have not been clearly identified. Abdominal computed tomography (CT) images of 723 patients which were taken between the years 2010 and 2014 were evaluated retrospectively. Patients with prior lumbar spinal surgery, serious congenital anomalies on CT, incomplete or complete lumbosacral transition, severe scoliosis, were excluded from the study. To eliminate the age- and sex-related differences in spinopelvic morphology, a study group was formed of the remaining subjects by including patients from a specific age group (30-35 yr) and same sex (females). For each patient the presence and grade of facet joint degeneration was investigated. In addition, pelvic incidence (PI), sacral slope and the angles of L1-L5 lumbar lordosis, sacral table, L5 vertebra posterior, and sacral kyphosis were measured for each patient. Sacral slope, sacral kyphosis, and L1-L5 lumbar lordosis angle were significantly higher in patients with osteoarthritic compared with normal subjects (P = 0.015, P = 0.018, P = 0.016). L5 vertebra posterior and sacral table angle were found to be significantly lower in patients with osteoarthritic than in normal subjects (P = 0.019, P = 0.007). The degree of facet joint degeneration was noticed to increase parallel to the decrease in the sacral table angle and L5 vertebra posterior angle, and to the increase in the L1-L5 lumbar lordosis, PI, and sacral slope. A close relation exists between the presence and degree of degeneration in the facet joint and lumbosacral pelvic morphology. Prevalence and degree of the degeneration in facet joint increases as the angle of sacral slope, L1-L5 lumbar lordosis, and PI increases or the angle of sacral table and L5 vertebra posterior decreases. 4.

  12. Optimized resolved rate control of seven-degree-of-freedom Laboratory Telerobotic Manipulator (LTM) with application to three-dimensional graphics simulation

    NASA Technical Reports Server (NTRS)

    Barker, L. Keith; Mckinney, William S., Jr.

    1989-01-01

    The Laboratory Telerobotic Manipulator (LTM) is a seven-degree-of-freedom robot arm. Two of the arms were delivered to Langley Research Center for ground-based research to assess the use of redundant degree-of-freedom robot arms in space operations. Resolved-rate control equations for the LTM are derived. The equations are based on a scheme developed at the Oak Ridge National Laboratory for computing optimized joint angle rates in real time. The optimized joint angle rates actually represent a trade-off, as the hand moves, between small rates (least-squares solution) and those rates which work toward satisfying a specified performance criterion of joint angles. In singularities where the optimization scheme cannot be applied, alternate control equations are devised. The equations developed were evaluated using a real-time computer simulation to control a 3-D graphics model of the LTM.

  13. LATDYN - PROGRAM FOR SIMULATION OF LARGE ANGLE TRANSIENT DYNAMICS OF FLEXIBLE AND RIGID STRUCTURES

    NASA Technical Reports Server (NTRS)

    Housner, J. M.

    1994-01-01

    LATDYN is a computer code for modeling the Large Angle Transient DYNamics of flexible articulating structures and mechanisms involving joints about which members rotate through large angles. LATDYN extends and brings together some of the aspects of Finite Element Structural Analysis, Multi-Body Dynamics, and Control System Analysis; three disciplines that have been historically separate. It combines significant portions of their distinct capabilities into one single analysis tool. The finite element formulation for flexible bodies in LATDYN extends the conventional finite element formulation by using a convected coordinate system for constructing the equation of motion. LATDYN's formulation allows for large displacements and rotations of finite elements subject to the restriction that deformations within each are small. Also, the finite element approach implemented in LATDYN provides a convergent path for checking solutions simply by increasing mesh density. For rigid bodies and joints LATDYN borrows extensively from methodology used in multi-body dynamics where rigid bodies may be defined and connected together through joints (hinges, ball, universal, sliders, etc.). Joints may be modeled either by constraints or by adding joint degrees of freedom. To eliminate error brought about by the separation of structural analysis and control analysis, LATDYN provides symbolic capabilities for modeling control systems which are integrated with the structural dynamic analysis itself. Its command language contains syntactical structures which perform symbolic operations which are also interfaced directly with the finite element structural model, bypassing the modal approximation. Thus, when the dynamic equations representing the structural model are integrated, the equations representing the control system are integrated along with them as a coupled system. This procedure also has the side benefit of enabling a dramatic simplification of the user interface for modeling control systems. Three FORTRAN computer programs, the LATDYN Program, the Preprocessor, and the Postprocessor, make up the collective LATDYN System. The Preprocessor translates user commands into a form which can be used while the LATDYN program provides the computational core. The Postprocessor allows the user to interactively plot and manage a database of LATDYN transient analysis results. It also includes special facilities for modeling control systems and for programming changes to the model which take place during analysis sequence. The documentation includes a Demonstration Problem Manual for the evaluation and verification of results and a Postprocessor guide. Because the program should be viewed as a byproduct of research on technology development, LATDYN's scope is limited. It does not have a wide library of finite elements, and 3-D Graphics are not available. Nevertheless, it does have a measure of "user friendliness". The LATDYN program was developed over a period of several years and was implemented on a CDC NOS/VE & Convex Unix computer. It is written in FORTRAN 77 and has a virtual memory requirement of 1.46 MB. The program was validated on a DEC MICROVAX operating under VMS 5.2.

  14. Use of the azimuthal resistivity technique for determination of regional azimuth of transmissivity

    USGS Publications Warehouse

    Carlson, D.

    2010-01-01

    Many bedrock units contain joint sets that commonly act as preferred paths for the movement of water, electrical charge, and possible contaminants associated with production or transit of crude oil or refined products. To facilitate the development of remediation programs, a need exists to reliably determine regional-scale properties of these joint sets: azimuth of transmissivity ellipse, dominant set, and trend(s). The surface azimuthal electrical resistivity survey method used for local in situ studies can be a noninvasive, reliable, efficient, and relatively cost-effective method for regional studies. The azimuthal resistivity survey method combines the use of standard resistivity equipment with a Wenner array rotated about a fixed center point, at selected degree intervals, which yields an apparent resistivity ellipse from which joint-set orientation can be determined. Regional application of the azimuthal survey method was tested at 17 sites in an approximately 500 km2 (193 mi2) area around Milwaukee, Wisconsin, with less than 15m (50 ft) overburden above the dolomite. Results of 26 azimuthal surveys were compared and determined to be consistent with the results of two other methods: direct observation of joint-set orientation and transmissivity ellipses from multiple-well-aquifer tests. The average of joint-set trend determined by azimuthal surveys is within 2.5?? of the average of joint-set trend determined by direct observation of major joint sets at 24 sites. The average of maximum of transmissivity trend determined by azimuthal surveys is within 5.7?? of the average of maximum of transmissivity trend determined for 14 multiple-well-aquifer tests. Copyright ?? 2010 The American Association of Petroleum Geologists/Division of Environmental Geosciences. All rights reserved.

  15. Near-field non-radial motion generation from underground chemical explosions in jointed granite

    NASA Astrophysics Data System (ADS)

    Vorobiev, Oleg; Ezzedine, Souheil; Hurley, Ryan

    2018-01-01

    This paper describes analysis of non-radial ground motion generated by chemical explosions in a jointed rock formation during the Source Physics Experiment (SPE). Such motion makes it difficult to discriminate between various subsurface events such as explosions, implosions (i.e. mine collapse) and earthquakes. We apply 3-D numerical simulations to understand experimental data collected during the SPEs. The joints are modelled explicitly as compliant thin inclusions embedded into the rock mass. Mechanical properties of the rock and the joints as well as the joint spacing and orientation are inferred from experimental test data, and geophysical and geological characterization of the SPE site which is dominantly Climax Stock granitic outcrop. The role of various factors characterizing the joints such as joint spacing, frictional properties, orientation and persistence in generation of non-radial motion is addressed. The joints in granite at the SPE site are oriented in nearly orthogonal directions with two vertical sets dipping at 70-80 degrees with the same strike angle, one vertical set almost orthogonal to the first two and one shallow angle joint set dipping 15 degrees. In this study we establish the relationship between the joint orientation and azimuthal variations in the polarity of the observed shear motion. The majority of the shear motion is generated due to the effects of non-elastic sliding on the joints near the source, where the wave can create significant shear stress to overcome the cohesive forces at the joints. Near the surface the joints are less confined and are subject to sliding when the pressure waves are reflected. In the far field, where the cohesive forces on the joints cannot be overcome, additional shear motion can be generated due to elastic anisotropy of the rock mass given by preferred spatial orientations of compliant joints.

  16. Ambulant adults with spastic cerebral palsy: the validity of lower limb joint angle measurements from sagittal video recordings.

    PubMed

    Larsen, Kerstin L; Maanum, Grethe; Frøslie, Kathrine F; Jahnsen, Reidun

    2012-02-01

    In the development of a clinical program for ambulant adults with cerebral palsy (CP), we investigated the validity of joint angles measured from sagittal video recordings and explored if movements in the transversal plane identified with three-dimensional gait analysis (3DGA) affected the validity of sagittal video joint angle measurements. Ten observers, and 10 persons with spastic CP (19-63 years), Gross Motor Function Classification System I-II, participated in the study. Concurrent criterion validity between video joint angle measurements and 3DGA was assessed by Bland-Altman plots with mean differences and 95% limits of agreement (LoA). Pearson's correlation coefficients (r) and scatter plots were used supplementary. Transversal kinematics ≥2 SD from our reference band were defined as increased movement in the transversal plane. The overall mean differences in degrees between joint angles measured by 3DGA and video recordings (3°, 5° and -7° for the hip, knee and ankle respectively) and corresponding LoA (18°, 10° and 15° for the hip, knee and ankle, respectively) demonstrated substantial discrepancies between the two methods. The correlations ranged from low (r=0.39) to moderate (r=0.68). Discrepancy between the two measurements was seen both among persons with and without the presence of deviating transversal kinematics. Quantifying lower limb joint angles from sagittal video recordings in ambulant adults with spastic CP demonstrated low validity, and should be conducted with caution. This gives implications for selecting evaluation method of gait. Copyright © 2011 Elsevier B.V. All rights reserved.

  17. A New Technique for Compensating Joint Limits in a Robot Manipulator

    NASA Technical Reports Server (NTRS)

    Litt, Jonathan; Hickman, Andre; Guo, Ten-Huei

    1996-01-01

    A new robust, optimal, adaptive technique for compensating rate and position limits in the joints of a six degree-of-freedom elbow manipulator is presented. In this new algorithm, the unmet demand as a result of actuator saturation is redistributed among the remaining unsaturated joints. The scheme is used to compensate for inadequate path planning, problems such as joint limiting, joint freezing, or even obstacle avoidance, where a desired position and orientation are not attainable due to an unrealizable joint command. Once a joint encounters a limit, supplemental commands are sent to other joints to best track, according to a selected criterion, the desired trajectory.

  18. The aeroacoustics of supersonic jets

    NASA Technical Reports Server (NTRS)

    Morris, Philip J.; McLaughlin, Dennis K.

    1995-01-01

    This research project was a joint experimental/computational study of noise in supersonic jets. The experiments were performed in a low to moderate Reynolds number anechoic supersonic jet facility. Computations have focused on the modeling of the effect of an external shroud on the generation and radiation of jet noise. This report summarizes the results of the research program in the form of the Masters and Doctoral theses of those students who obtained their degrees with the assistance of this research grant. In addition, the presentations and publications made by the principal investigators and the research students is appended.

  19. Quick-connect threaded attachment joint

    NASA Technical Reports Server (NTRS)

    Lucy, M. H.; Messick, W. R.; Vasquez, P.

    1979-01-01

    Joint is self-aligning and tightens with only sixty-five degrees of rotation for quick connects and disconnects. Made of injection-molded plastics or cast or machined aluminum, joint can carry wires, tubes, liquids, or gases. When two parts of joint are brought together, their shapes align them. Small projections on male section and slots on female section further aid alignment; slight rotation of male form engages projections in slots. At this point, threads engage and male section is rotated until joint is fully engaged.

  20. Experiences in the development of rotary joints for robotic manipulators in space applications

    NASA Technical Reports Server (NTRS)

    Priesett, Klaus

    1992-01-01

    European developments in robotics for space applications have resulted in human arm-like manipulators with six or more rotational degrees of freedom. The rotary joints including their own electromechanical actuator and feedback sensors must be very compact units. The specific joint concept is presented as evolved so far. The problems encountered during the first hardware development phases are covered on both component and joint level.

  1. Position of the prosthesis and the incidence of dislocation following total hip replacement.

    PubMed

    He, Rong-xin; Yan, Shi-gui; Wu, Li-dong; Wang, Xiang-hua; Dai, Xue-song

    2007-07-05

    Dislocation is the second most common complication of hip replacement surgery, and impact of the prosthesis is believed to be the fundamental reason. The present study employed Solidworks 2003 and MSC-Nastran software to analyze the three dimensional variables in order to investigate how to prevent dislocation following hip replacement surgery. Computed tomography (CT) imaging was used to collect femoral outline data and Solidworks 2003 software was used to construct the cup model with variabilities. Nastran software was used to evaluate dislocation at different prosthesis positions and different geometrical shapes. Three dimensional movement and results from finite element method were analyzed and the values of dislocation resistance index (DRI), range of motion to impingement (ROM-I), range of motion to dislocation (ROM-D) and peak resisting moment (PRM) were determined. Computer simulation was used to evaluate the range of motion of the hip joint at different prosthesis positions. Finite element analysis showed: (1) Increasing the ratio of head/neck increased the ROM-I values and moderately increased ROM-D and PRM values. Increasing the head size significantly increased PRM and to some extent ROM-I and ROM-D values, which suggested that there would be a greater likelihood of dislocation. (2) Increasing the anteversion angle increased the ROM-I, ROM-D, PRM, energy required for dislocation (ENERGY-D) and DRI values, which would increase the stability of the joint. (3) As the chamber angle was increased, ROM-I, ROM-D, PRM, Energy-D and DRI values were increased, resulting in improved joint stability. Chamber angles exceeding 55 degrees resulted in increases in ROM-I and ROM-D values, but decreases in PRM, Energy-D, and DRI values, which, in turn, increased the likelihood of dislocation. (4) The cup, which was reduced posteriorly, reduced ROM-I values (2.1 -- 5.3 degrees ) and increased the DRI value (0.073). This suggested that the posterior high side had the effect of 10 degrees anteversion angle. Increasing the head/neck ratio increases joint stability. Posterior high side reduced the range of motion of the joint but increased joint stability; Increasing the anteversion angle increases DRI values and thus improve joint stability; Increasing the chamber angle increases DRI values and improves joint stability. However, at angles exceeding 55 degrees , further increases in the chamber angle result in decreased DRI values and reduce the stability of the joint.

  2. Program Calculates Forces in Bolted Structural Joints

    NASA Technical Reports Server (NTRS)

    Buder, Daniel A.

    2005-01-01

    FORTRAN 77 computer program calculates forces in bolts in the joints of structures. This program is used in conjunction with the NASTRAN finite-element structural-analysis program. A mathematical model of a structure is first created by approximating its load-bearing members with representative finite elements, then NASTRAN calculates the forces and moments that each finite element contributes to grid points located throughout the structure. The user selects the finite elements that correspond to structural members that contribute loads to the joints of interest, and identifies the grid point nearest to each such joint. This program reads the pertinent NASTRAN output, combines the forces and moments from the contributing elements to determine the resultant force and moment acting at each proximate grid point, then transforms the forces and moments from these grid points to the centroids of the affected joints. Then the program uses these joint loads to obtain the axial and shear forces in the individual bolts. The program identifies which bolts bear the greatest axial and/or shear loads. The program also performs a fail-safe analysis in which the foregoing calculations are repeated for a sequence of cases in which each fastener, in turn, is assumed not to transmit an axial force.

  3. 76 FR 35474 - UAW-Chrysler Technical Training Center, Technology Training Joint Programs Staff, Including On...

    Federal Register 2010, 2011, 2012, 2013, 2014

    2011-06-17

    ...-Chrysler Technical Training Center, Technology Training Joint Programs Staff, Including On-Site Leased Workers From Cranks, O/E Learning, DBSI, IDEA, and Tonic/MVP, Detroit, MI; UAW-Chrysler Technical Training... workers and former workers of UAW-Chrysler Technical Training Center, Technology Training Joint Programs...

  4. Training Engineers of Joint Programs for the European Aerospace Industry.

    ERIC Educational Resources Information Center

    Thomas, Jurgen

    1985-01-01

    Examines topics and issues related to training engineers of joint programs for the European aerospace industry. Forms of cooperation, European educational systems, and skills needed to successfully work as an engineer in a joint program for the European aircraft industry are the major areas addressed. (JN)

  5. Characteristics of upper limb muscular strength in male wheelchair tennis players

    PubMed Central

    Moon, Hyo-Bin; Park, Seung-Jae; Kim, Al-Chan; Jang, Jee-Hun

    2013-01-01

    The purpose of this study was to identify the characteristics of muscular strength in upper limb and to present the preliminary information for development of sports injury prevention program and exercise rehabilitation program in wheelchair tennis players. Participants were 12 male wheelchair tennis players. Muscular strength was measured in shoulder and elbow joints with isokinetic dynamometer. Ipsilateral (IR) and bilateral (BR) balance ratio were calculated with isokinetic strength at 60°/sec. As a result, extension strength (ES) was significantly higher than flexion strength (FS) (P< 0.001), and IR in both sides and BR in ES were maintained within normal range whereas BR in FS was lower than normal range in shoulder joint. In elbow joint FS was significantly higher than ES (P< 0.05), and IR and BR were lower than normal range. Consequently, the different tendency in IR between shoulder and elbow joints and lower IR and BR in elbow joints could be the characteristics in male wheelchair tennis players. It is suggested that flexor strengthening program in nondominant shoulder joint, extensor strengthening program in both elbow joint, and flexor strengthening program in non-dominant elbow joint should be introduced for male wheelchair tennis players. PMID:24278887

  6. A Unique Partnership to Promote Diversity in the Geosciences, San Jose, California

    NASA Astrophysics Data System (ADS)

    Sedlock, R.; Metzger, E.; Johnson, D.

    2006-12-01

    We report here on a particularly satisfying partnership of academic institutions that focuses on enhancing the participation of underrepresented students in the geosciences. The Bay Area Earth Science Institute (BAESI) at San José State University (SJSU) has provided professional development opportunities to over 1,500 area teachers since 1990. BAESI offerings include summer and weekend workshops, field trips, classroom visits, and a lending library of curricula, sample sets, A/V materials, and equipment. The National Hispanic University (NHU) is a private, non-profit university that enrolls about 700 students, 80% of whom are of Hispanic descent. Another 13% are from other minority groups, 74% are from low-income families, and 70% are women. NHU houses the Latino College Preparatory Academy (LCPA), a charter high school that provides an alternative for students who struggle in traditional schools due to language issues. In the 1990s, administrators at SJSU and NHU set up formal agreements about course articulation, reciprocity, and joint degree programs. In 2002, informal discussions between BAESI and NHU staff led to collaboration on an NSF proposal to strengthen NHU's geoscience curriculum. Since then, the scope of BAESI-NHU actions has expanded greatly: (1) NHU and LCPA staff attended a week-long BAESI professional development workshop funded by NSF, and have attended numerous BAESI field trips. (2) BAESI staff visit NHU and LCPA classrooms to showcase SJSU's Geology Department and to enrich existing Chemistry and Physics classes with geoscience applications. (3) A nascent "Geologist-In-Residence" program pairs SJSU geology students with teachers at LCPA. (4) NHU students have interned with Metzger on local research projects. (5) BAESI brokered donation of an extensive USGS rock collection to NHU. (6) NHU, BAESI, and NASA-Ames staff collaborate on an online Earth Science curriculum for middle-school teachers. (7) We will adapt BAESI summer workshops to a one-week course in effective teaching of high-school science that will be taught during intersession in NHU's Teacher Education Department. We have recently received funding for a collaborative project from NSF's Geoscience Education program to create a joint degree program wherein NHU offers the lower division coursework and bestows an A.S. degree in mathematics and science with geoscience emphasis, and SJSU offers the upper-division coursework and the B.S. degree in geoscience. Our collaborations focus on providing teachers with professional development and educational resources to help underrepresented students receive quality instruction in the geosciences. Participation of NHU teachers- in-training provides a long-term means for spreading quality geoscience teaching to precollege classrooms throughout Santa Clara County, including the largely minority classrooms that NHU teachers are specially trained to staff.

  7. MR imaging of patients with lateral epicondylitis of the elbow: is the common extensor tendon an isolated lesion?

    PubMed

    Qi, Liang; Zhu, Zheng-Feng; Li, Feng; Wang, Ren-Fa

    2013-01-01

    To investigate whether an injury of the common extensor tendon (CET) is associated with other abnormalities in the elbow joint and find the potential relationships between these imaging features by using a high-resolution magnetic resonance imaging (MRI). Twenty-three patients were examined with 3.0 T MR. Two reviewers were recruited for MR images evaluation. Image features were recorded in terms of (1) the injury degree of CET; (2) associated injuries in the elbow joint. Spearman's rank correlation analysis was performed to analyze the relationships between the injury degree of CET and associated abnormalities of the elbow joint, correlations were considered significant at p<0.05. Total 24 elbows in 23 patients were included. Various degrees of injuries were found in total 24 CETs (10 mild, 7 moderate and 7 severe). Associated abnormalities were detected in accompaniments of the elbow joints including ligaments, tendons, saccussynovialis and muscles. A significantly positive correlation (r = 0.877,p<0.01) was found in injuries of CET and lateral ulnar collateral ligament (LUCL). Injury of the CET is not an isolated lesion for lateral picondylitis, which is mostly accompanied with other abnormalities, of which the LUCL injury is the most commonly seen in lateral epicondylitis, and there is a positive correlation between the injury degree in CET and LUCL.

  8. Six-degree-of-freedom parallel minimanipulator with three inextensible limbs

    NASA Technical Reports Server (NTRS)

    Tahmasebi, Farhad (Inventor); Tsai, Lung-Wen (Inventor)

    1994-01-01

    A Six-Degree-of-Freedom Parallel-Manipulator having three inextensible limbs for manipulating a platform is described. The three inextensible limbs are attached via universal joints to the platform at non-collinear points. Each of the inextensible limbs is also attached via universal joints to a two-degree-of-freedom parallel driver such as a five-bar linkage, a pantograph, or a bidirectional linear stepper motor. The drivers move the lower ends of the limbs parallel to a fixed base and thereby provide manipulation of the platform. The actuators are mounted on the fixed base without using any power transmission devices such as gears or belts.

  9. 7 CFR 3405.7 - Joint project proposals.

    Code of Federal Regulations, 2011 CFR

    2011-01-01

    ... 7 Agriculture 15 2011-01-01 2011-01-01 false Joint project proposals. 3405.7 Section 3405.7... AGRICULTURE HIGHER EDUCATION CHALLENGE GRANTS PROGRAM Program Description § 3405.7 Joint project proposals. Applicants are encouraged to submit joint project proposals as defined in § 3405.2(m), which address regional...

  10. Adhesive-bonded scarf and stepped-lap joints

    NASA Technical Reports Server (NTRS)

    Hart-Smith, L. J.

    1973-01-01

    Continuum mechanics solutions are derived for the static load-carrying capacity of scarf and stepped-lap adhesive-bonded joints. The analyses account for adhesive plasticity and adherend stiffness imbalance and thermal mismatch. The scarf joint solutions include a simple algebraic formula which serves as a close lower bound, within a small fraction of a per cent of the true answer for most practical geometries and materials. Digital computer programs were developed and, for the stepped-lap joints, the critical adherend and adhesive stresses are computed for each step. The scarf joint solutions exhibit grossly different behavior from that for double-lap joints for long overlaps inasmuch as that the potential bond shear strength continues to increase with indefinitely long overlaps on the scarf joints. The stepped-lap joint solutions exhibit some characteristics of both the scarf and double-lap joints. The stepped-lap computer program handles arbitrary (different) step lengths and thickness and the solutions obtained have clarified potentially weak design details and the remedies. The program has been used effectively to optimize the joint proportions.

  11. The Devil in the Detail: Contradictory National Requirements and Bologna Master Degrees

    ERIC Educational Resources Information Center

    Sin, Cristina

    2013-01-01

    This article compares the national-level requirements for master degree provision in England, Denmark and Portugal following the implementation of the Bologna Process, and ponders upon the reconcilability of these requirements in cross-national initiatives (e.g. joint degrees). In all three countries, master degrees have to comply with the…

  12. SRB Environment Evaluation and Analysis. Volume 2: RSRB Joint Filling Test/Analysis Improvements

    NASA Technical Reports Server (NTRS)

    Knox, E. C.; Woods, G. Hamilton

    1991-01-01

    Following the Challenger accident a very comprehensive solid rocket booster (SRB) redesign program was initiated. One objective of the program was to develop expertise at NASA/MSFC in the techniques for analyzing the flow of hot gases in the SRB joints. Several test programs were undertaken to provide a data base of joint performance with manufactured defects in the joints to allow hot gases to fill the joints. This data base was used also to develop the analytical techniques. Some of the test programs were Joint Environment Simulator (JES), Nozzle Joint Environment Simulator (NJES), Transient Pressure Test Article (TPTA), and Seventy-Pound Charge (SPC). In 1988 the TPTA test hardware was moved from the Utah site to MSFC and several RSRM tests were scheduled, to be followed by tests for the ASRM program. REMTECH Inc. supported these activities with pretest estimates of the flow conditions in the test joints, and post-test analysis and evaluation of the measurements. During this support REMTECH identified deficiencies in the gas-measurement instrumentation that existed in the TPTA hardware, made recommendations for its replacement, and identified improvements to the analytical tools used in the test support. Only one test was completed under the TPTA RSRM test program, and those scheduled for the ASRM were rescheduled to a time after the expiration of this contract. The attention of this effort was directed toward improvements in the analytical techniques in preparation for when the ASRM program begins.

  13. Reducing the Volume of NASA Earth-Science Data

    NASA Technical Reports Server (NTRS)

    Lee, Seungwon; Braverman, Amy J.; Guillaume, Alexandre

    2010-01-01

    A computer program reduces data generated by NASA Earth-science missions into representative clusters characterized by centroids and membership information, thereby reducing the large volume of data to a level more amenable to analysis. The program effects an autonomous data-reduction/clustering process to produce a representative distribution and joint relationships of the data, without assuming a specific type of distribution and relationship and without resorting to domain-specific knowledge about the data. The program implements a combination of a data-reduction algorithm known as the entropy-constrained vector quantization (ECVQ) and an optimization algorithm known as the differential evolution (DE). The combination of algorithms generates the Pareto front of clustering solutions that presents the compromise between the quality of the reduced data and the degree of reduction. Similar prior data-reduction computer programs utilize only a clustering algorithm, the parameters of which are tuned manually by users. In the present program, autonomous optimization of the parameters by means of the DE supplants the manual tuning of the parameters. Thus, the program determines the best set of clustering solutions without human intervention.

  14. One Mission-Centered, Market-Smart Globalization Response: A Case Study of the Georgia Tech-Emory University Biomedical Engineering Curricular Joint Venture

    ERIC Educational Resources Information Center

    Burriss, Annie Hunt

    2010-01-01

    One innovative, higher-education response to globalization and changing fiscal realities is the curricular joint venture (CJV), a formal collaboration between academic institutions that leverages missions through new joint degrees and research not previously offered by collaborating institutions (Eckel, 2003). In 1997, a pioneering biomedical…

  15. Rehabilitation of children and infants with osteogenesis imperfecta. A program for ambulation.

    PubMed

    Gerber, L H; Binder, H; Weintrob, J; Grange, D K; Shapiro, J; Fromherz, W; Berry, R; Conway, A; Nason, S; Marini, J

    1990-02-01

    Management of children and infants with osteogenesis imperfecta (OI) poses difficult decisions for pediatricians, orthopedists, and physiatrists. These children are frequently frail with disabling bone and joint deformities and fractures. In an eight-year cumulative management of 12 children with OI, a comprehensive program included strengthening exercises to the pelvic girdle and lower extremity muscles, in addition to pool exercises and molded seating to support upright posture. Long leg braces were fitted when the children were able to sit unsupported. All 12 were fitted with braces; nine were functional ambulators, and three were home ambulators. Six children required femoral plating or rodding, two of whom subsequently had the metal removed. Lower extremity fractures averaged one and one-half per year prior to bracing for nine children who had fractures. There was 0.83 fracture per year for the ten children who had fractures after bracing. The degree of femoral bowing increased in four, decreased in four, and remained unchanged in four, while the degree of tibial bowing increased in two, decreased in nine, and remained unchanged in one during the observation period. A comprehensive rehabilitation program and long leg bracing with surgical operations on the femur result in a high level of functional activity for children with OI with an acceptable level of risk for fracture.

  16. 25 CFR 20.505 - What services are provided jointly with the Child Assistance Program?

    Code of Federal Regulations, 2010 CFR

    2010-04-01

    ... 25 Indians 1 2010-04-01 2010-04-01 false What services are provided jointly with the Child... SERVICES FINANCIAL ASSISTANCE AND SOCIAL SERVICES PROGRAMS Child Assistance How Child Assistance Funds Can Be Used § 20.505 What services are provided jointly with the Child Assistance Program? The services...

  17. Public health education at the University of California, Davis: past, present, and future programs.

    PubMed

    Hird, David W; Lloyd, K C Kent; McCurdy, Stephen A; Schenker, Marc B; Troidl, John J; Kass, Philip H

    2008-01-01

    This article reviews the history of public-health education at the University of California, Davis, from the inception of the Master of Preventive Veterinary Medicine Program in the School of Veterinary Medicine through the creation of the Master of Public Health Program offered jointly by the Schools of Medicine and Veterinary Medicine. The long history of collaborative teaching and research between the schools, as well as the university's close proximity to and relationship with numerous university-affiliated and state public-health agencies, has created remarkable opportunities for novel and creative public-health education. The university is already anticipating the approval of a School of Public Health on its campus, which will create even more educational opportunities in both human and veterinary public-health disciplines. Given the projected shortfall of veterinarians entering such fields, the opportunity of a novel Doctor of Public Health degree program specifically suited to the needs of veterinary medicine is also discussed as a means of addressing this shortage.

  18. 21 CFR 888.6 - Degree of constraint.

    Code of Federal Regulations, 2013 CFR

    2013-04-01

    ... ORTHOPEDIC DEVICES General Provisions § 888.6 Degree of constraint. Certain joint prostheses provide more... affecting the safety and effectiveness of orthopedic prostheses. FDA is defining the following standard...

  19. 21 CFR 888.6 - Degree of constraint.

    Code of Federal Regulations, 2011 CFR

    2011-04-01

    ... ORTHOPEDIC DEVICES General Provisions § 888.6 Degree of constraint. Certain joint prostheses provide more... affecting the safety and effectiveness of orthopedic prostheses. FDA is defining the following standard...

  20. 21 CFR 888.6 - Degree of constraint.

    Code of Federal Regulations, 2014 CFR

    2014-04-01

    ... ORTHOPEDIC DEVICES General Provisions § 888.6 Degree of constraint. Certain joint prostheses provide more... affecting the safety and effectiveness of orthopedic prostheses. FDA is defining the following standard...

  1. 21 CFR 888.6 - Degree of constraint.

    Code of Federal Regulations, 2012 CFR

    2012-04-01

    ... ORTHOPEDIC DEVICES General Provisions § 888.6 Degree of constraint. Certain joint prostheses provide more... affecting the safety and effectiveness of orthopedic prostheses. FDA is defining the following standard...

  2. [Strategies for simultaneous control of the equilibrium and of the head position during the raising movement of a leg].

    PubMed

    Mouchnino, L; Aurenty, R; Massion, J; Pedotti, A

    1991-01-01

    The coordination between equilibrium control and the ability to maintain the position of given segments (head, trunk) was studied in standing subjects, instructed to raise one leg laterally at an angle of 45 degrees in response to a light. Two sources of light placed at eye level indicated the side on which the movement was to be performed. Two populations were compared: naive subjects and dancers. Two control strategies were identified. An "inclination" strategy was used by the naive subjects. This consisted of an external rotation of the body around the antero-posterior ankle joint axis; a counter-rotation of the head with respect to the trunk was observed, which ensured some stabilization in the horizontal plane of the interorbital line. A "translation" strategy was used by the dancers. Here the external rotation of the leg around the ankle joint was associated with a feed-forward counter-rotation of the trunk around the coxofemoral joint so that the horizontality of the interorbital line and the verticality of the trunk axis were maintained. This new coordination results from a long-term training and indicates that a new motor program has been elaborated.

  3. MRI evaluation of infectious and non-infectious synovitis: preliminary studies in a rabbit model.

    PubMed

    Strouse, P J; Londy, F; DiPietro, M A; Teo, E L; Chrisp, C E; Doi, K

    1999-05-01

    Literature on magnetic resonance imaging (MR) evaluation of inflammatory joint effusions is sparse. To describe an animal model for studying infectious and non-infectious joint effusions with magnetic resonance imaging. Ten rabbit knees with septic arthritis and four with talc synovitis were imaged with MR. Contralateral knees injected with saline served as controls. Fat saturation T2-weighted and gadolinium-enhanced T1-weighted images were assessed for joint effusion, and periarticular and adjacent intraosseous increased signal or enhancement. Each knee was cultured and underwent pathologic examination. Both Staphylococcus aureus and talc produced effusions in all knees. The degree of periarticular signal and enhancement was greater in infected knees than talc-injected knees. No abnormal enhancement was seen within bone. Pathologic examination showed a greater degree of inflammation and joint destruction in the infected knees, but no evidence of osteomyelitis. A greater degree of abnormal signal and enhancement seen on MR suggests a more vigorous inflammatory process, as seen with septic arthritis. In spite of advanced septic arthritis, no enhancement was evident within bone, suggesting that enhancement within bone is not an expected finding in isolated septic arthritis and should raise concern for osteomyelitis.

  4. Identification of Nonlinear Micron-Level Mechanics for a Precision Deployable Joint

    NASA Technical Reports Server (NTRS)

    Bullock, S. J.; Peterson, L. D.

    1994-01-01

    The experimental identification of micron-level nonlinear joint mechanics and dynamics for a pin-clevis joint used in a precision, adaptive, deployable space structure are investigated. The force-state mapping method is used to identify the behavior of the joint under a preload. The results of applying a single tension-compression cycle to the joint under a tensile preload are presented. The observed micron-level behavior is highly nonlinear and involves all six rigid body motion degrees-of-freedom of the joint. it is also suggests that at micron levels of motion modelling of the joint mechanics and dynamics must include the interactions between all internal components, such as the pin, bushings, and the joint node.

  5. [Application and effect of auricular acupoint pressing for analgesia in perioperative period of total knee joint replacement].

    PubMed

    Tong, Pei-Jian; Wang, Hai-Dong; Ma, Zhen-Chuan

    2010-09-01

    To observe the effect of auricular acupoint pressing (AAP) for analgesia during perioperative period of total knee joint replacement. Sixty patients with osteoarthritis of ASA grade I - III scheduled to receive unilateral total knee joint replacement were equally randomized into the AAP group and the control group, 30 in each group. The general anesthesia on all patients was implemented by physicians of an identical group through endotracheal intubation. To the patients in the AAP group, AAP with Vaccaria seed was applied before operation, and the local analgesia on affected limb with acupoint pasting was used after operation. Besides, administering of celecoxib 400 mg on the day before operation, and celecoxib 200 mg twice daily post-operation was given to all patients. When the visual analogue scales (VAS) reached more than 7 points, 0.1 g of bucinnazine hydrochloride was given for supplement. Meantime, same post-operative training methods were adopted in both groups. The resting VAS pain scores, contentment of sedation, incidence of adverse event, postoperative range of motion (ROM) of knee joint and Hospital for Special Surgery (HSS) score were recorded. The resting VAS pain scores at 6 h and 24 h after operation was 5.99 +/- 0.67 scores and 4.26 +/- 0.59 scores in the AAP group respectively, which was significantly lower than that in the control group at the corresponding time (7.02 +/- 0.85 scores and 4.92 +/- 0.43 scores, P < 0.01); but it showed insignificant difference between the two groups at 1 h and 48 h after operation (P > 0.05); sedation contentment in the two groups was similar; incidence of adverse event in the AAP groups seemed lower (4 cases vs. 11 cases), but the intergroup difference was statistically insignificant (P > 0.05). ROM before surgery were 75.63 degrees +/- 5.74 degrees and 75.43 degrees +/- 5.63 degrees in the two groups respectively, showing no significant difference (P > 0.05), two weeks after operation, the initiative ROM raised to 96.50 degrees +/- 3.79 degrees and 93.50 degrees +/- 3.50 degrees, and the passive ROM reached 107.8 degrees +/- 3.37 degrees and 105.27 degrees +/- 3.25 degrees in the two groups respectively, with statistical significance between them (P < 0.05). HSS score was similar between groups before operation (60.23 +/- 3.44 scores vs. 61.70 +/- 2.83 scores, P > 0.05); while it became 86.97 +/- 2.33 scores and 85.37 +/- 2.30 scores after operation, showing significant difference between groups (P < 0.05). Applying auricular acupoint pressing in perioperative period of knee joint replacement is favorable for alleviating postoperative pain, decreasing narcotic consumption, and promoting early rehabilitation, and it has the advantages of low cost, less complication, simple manipulation and high safety.

  6. Defense Acquisitions: Assessments of Selected Weapon Programs

    DTIC Science & Technology

    2016-03-01

    Increment 3 81 Indirect Fire Protection Capability Increment 2-Intercept Block 1 (IFPC Inc 2-I Block 1) 83 Improved Turbine Engine Program (ITEP...ITEP Improved Turbine Engine Program JAGM Joint Air-to-Ground Missile JLTV Joint Light Tactical Vehicle JSTARS Recap Joint Surveillance Target...Attack Radar System Recap 09/2017 —-   Improved Turbine Engine Program 06/2018 O O O Amphibious Ship Replacement 09/2018 O O Advanced Pilot

  7. A practical solution to reduce soft tissue artifact error at the knee using adaptive kinematic constraints.

    PubMed

    Potvin, Brigitte M; Shourijeh, Mohammad S; Smale, Kenneth B; Benoit, Daniel L

    2017-09-06

    Musculoskeletal modeling and simulations have vast potential in clinical and research fields, but face various challenges in representing the complexities of the human body. Soft tissue artifact from skin-mounted markers may lead to non-physiological representation of joint motions being used as inputs to models in simulations. To address this, we have developed adaptive joint constraints on five of the six degree of freedom of the knee joint based on in vivo tibiofemoral joint motions recorded during walking, hopping and cutting motions from subjects instrumented with intra-cortical pins inserted into their tibia and femur. The constraint boundaries vary as a function of knee flexion angle and were tested on four whole-body models including four to six knee degrees of freedom. A musculoskeletal model developed in OpenSim simulation software was constrained to these in vivo boundaries during level gait and inverse kinematics and dynamics were then resolved. Statistical parametric mapping indicated significant differences (p<0.05) in kinematics between bone pin constrained and unconstrained model conditions, notably in knee translations, while hip and ankle flexion/extension angles were also affected, indicating the error at the knee propagates to surrounding joints. These changes to hip, knee, and ankle kinematics led to measurable changes in hip and knee transverse plane moments, and knee frontal plane moments and forces. Since knee flexion angle can be validly represented using skin mounted markers, our tool uses this reliable measure to guide the five other degrees of freedom at the knee and provide a more valid representation of the kinematics for these degrees of freedom. Copyright © 2017 Elsevier Ltd. All rights reserved.

  8. MIXOR: a computer program for mixed-effects ordinal regression analysis.

    PubMed

    Hedeker, D; Gibbons, R D

    1996-03-01

    MIXOR provides maximum marginal likelihood estimates for mixed-effects ordinal probit, logistic, and complementary log-log regression models. These models can be used for analysis of dichotomous and ordinal outcomes from either a clustered or longitudinal design. For clustered data, the mixed-effects model assumes that data within clusters are dependent. The degree of dependency is jointly estimated with the usual model parameters, thus adjusting for dependence resulting from clustering of the data. Similarly, for longitudinal data, the mixed-effects approach can allow for individual-varying intercepts and slopes across time, and can estimate the degree to which these time-related effects vary in the population of individuals. MIXOR uses marginal maximum likelihood estimation, utilizing a Fisher-scoring solution. For the scoring solution, the Cholesky factor of the random-effects variance-covariance matrix is estimated, along with the effects of model covariates. Examples illustrating usage and features of MIXOR are provided.

  9. Resistance Exercise for Knee Osteoarthritis

    PubMed Central

    Vincent, Kevin R.; Vincent, Heather K.

    2013-01-01

    The initiation, progression, and severity of knee osteoarthritis (OA) has been associated with decreased muscular strength and alterations in joint biomechanics. Chronic OA pain may lead to anxiety, depression, fear of movement, and poor psychological outlook. The fear of movement may prevent participation in exercise and social events which could lead to further physical and social isolation. Resistance exercise (RX) has been shown to be an effective intervention both for decreasing pain and for improving physical function and self-efficacy. RX may restore muscle strength and joint mechanics while improving physical function. RX may also normalize muscle firing patterns and joint biomechanics leading to reductions in joint pain and cartilage degradation. These physical adaptations could lead to improved self-efficacy and decreased anxiety and depression. RX can be prescribed and performed by patients across the OA severity spectrum. When designing and implementing an RX program for a patient with knee OA, it is important to consider both the degree of OA severity as well as the level of pain. RX, either in the home or at a fitness facility, is an important component of a comprehensive regimen designed to offset the physical and psychological limitations associated with knee OA. Unique considerations for this population include: 1) monitoring pain during and after exercise, 2) providing days of rest when disease flares occur, and 3) infusing variety into the exercise regimen to encourage adherence. PMID:22632702

  10. 75 FR 6250 - ITS Joint Program Office; Intelligent Transportation Systems Program Advisory Committee; Notice...

    Federal Register 2010, 2011, 2012, 2013, 2014

    2010-02-08

    ... DEPARTMENT OF TRANSPORTATION ITS Joint Program Office; Intelligent Transportation Systems Program... the Intelligent Transportation Systems (ITS) Program Advisory Committee (ITSPAC). The Web conference... Transportation on all matters relating to the study, development, and implementation of intelligent...

  11. An activated energy approach for accelerated testing of the deformation of UHMWPE in artificial joints.

    PubMed

    Galetz, Mathias Christian; Glatzel, Uwe

    2010-05-01

    The deformation behavior of ultrahigh molecular polyethylene (UHMWPE) is studied in the temperature range of 23-80 degrees C. Samples are examined in quasi-static compression, tensile and creep tests to determine the accelerated deformation of UHMWPE at elevated temperatures. The deformation mechanisms under compression load can be described by one strain rate and temperature dependent Eyring process. The activation energy and volume of that process do not change between 23 degrees C and 50 degrees C. This suggests that the deformation mechanism under compression remains stable within this temperature range. Tribological tests are conducted to transfer this activated energy approach to the deformation behavior under loading typical for artificial knee joints. While this approach does not cover the wear mechanisms close to the surface, testing at higher temperatures is shown to have a significant potential to reduce the testing time for lifetime predictions in terms of the macroscopic creep and deformation behavior of artificial joints. Copyright 2010. Published by Elsevier Ltd.

  12. [Conservative treatment of metacarpal fracture].

    PubMed

    Prokop, A; Helling, H J; Kulus, S; Rehm, K E

    2002-01-01

    Conservative treatment of meatacarpale fracture is recommended if there are no joint displacement, rotation failures, displacement over 30 degrees ad axim and shortening over 5 mm. Operative procedures should be done in open fractures and serial of fractures of metacarpale bones. Early functionally treatment should be done in stable, not displaced fractures. Cast can be used only for a short time in full extended position of fingers and flexion in metacarpo-phalangeal joint in 60-90 degrees. Twin-tapes after reduction of edema allowed free range of motion by fixed rotation. Closed reduction of displaced fractures of fifth metacarpal bone (boxer's fracture) isn't successful. Cases with displacement over 30 degrees may be operatively treated by intramedullary stabilization.

  13. [Treatment of metacarpal fractures].

    PubMed

    Prokop, A; Jubel, A; Helling, H J; Kulus, S; Rehm, K E

    2002-09-01

    Conservative treatment of metacarpal fractures is recommended if there is no joint displacement, malrotation, displacement of over 30 degrees ad axim and shortening of over 5 mm. Surgery should be performed in open fractures and serial fractures of metacarpal bone. Early functional treatment should be carried out in stable, not displaced fractures. A cast can be used for a short period in full extended position of fingers and flexion in metacarpo-phalangeal joint in 60 - 90 degrees. Twin-tape fixation allows functional treatment after soft-tissue swelling has disappeared. Closed reduction of displaced fractures of the fifth metacarpal bone (boxer's fracture) is not successful. Cases with displacement of over 30 degrees may be treated surgically by intramedullary stabilisation.

  14. Dynamic motion planning of 3D human locomotion using gradient-based optimization.

    PubMed

    Kim, Hyung Joo; Wang, Qian; Rahmatalla, Salam; Swan, Colby C; Arora, Jasbir S; Abdel-Malek, Karim; Assouline, Jose G

    2008-06-01

    Since humans can walk with an infinite variety of postures and limb movements, there is no unique solution to the modeling problem to predict human gait motions. Accordingly, we test herein the hypothesis that the redundancy of human walking mechanisms makes solving for human joint profiles and force time histories an indeterminate problem best solved by inverse dynamics and optimization methods. A new optimization-based human-modeling framework is thus described for predicting three-dimensional human gait motions on level and inclined planes. The basic unknowns in the framework are the joint motion time histories of a 25-degree-of-freedom human model and its six global degrees of freedom. The joint motion histories are calculated by minimizing an objective function such as deviation of the trunk from upright posture that relates to the human model's performance. A variety of important constraints are imposed on the optimization problem, including (1) satisfaction of dynamic equilibrium equations by requiring the model's zero moment point (ZMP) to lie within the instantaneous geometrical base of support, (2) foot collision avoidance, (3) limits on ground-foot friction, and (4) vanishing yawing moment. Analytical forms of objective and constraint functions are presented and discussed for the proposed human-modeling framework in which the resulting optimization problems are solved using gradient-based mathematical programming techniques. When the framework is applied to the modeling of bipedal locomotion on level and inclined planes, acyclic human walking motions that are smooth and realistic as opposed to less natural robotic motions are obtained. The aspects of the modeling framework requiring further investigation and refinement, as well as potential applications of the framework in biomechanics, are discussed.

  15. Modified Essex-Lopresti / Westheus reduction for displaced intra-articular fractures of the calcaneus. Description of surgical technique and early outcomes.

    PubMed

    Pillai, Anand; Basappa, Prabhudeva; Ehrendorfer, Stefan

    2007-02-01

    We describe a modification of the classical Essex-Lopresti manoeuvre for the indirect reduction and stabilisation of displaced intra-articular fractures of the calcaneus. The radiological and functional results achieved using this technique in 15 patients is presented. Ten tongue-shaped and 8 joint depression type fractures were treated by the new method, incorporating the use of an additional traction pin. The pre and postoperative Böhler angles as well as the correction achieved were documented. Functional assessment was carried out using the Maryland Foot Score. The mean pre-operative Böhler angle in the joint depression group was 5.5 degrees, and in the tongue shaped fracture group 5 degrees. The mean postoperative Böhler angle in the joint depression group was 15.8 degrees, and in the tongue shape group was 23.25 degrees. At a mean follow-up of 28 months the joint depression group scored 51/100 on the foot score, and the tongue shaped fracture group 77/100. The mean correction achieved as well as the mean overall functional scores were significantly better in the tongue shaped group. The technique described has much promise in the management of selected displaced intra-articular fractures of the calcaneus (true tongue shaped / Sanders II), and may also have a limited role in other fracture types in patients with significant co-morbidities, soft tissue compromise and poor healing potential.

  16. Characteristics of clinical shoulder research over the last decade: a review of shoulder articles in The Journal of Bone & Joint Surgery from 2004 to 2014.

    PubMed

    Gartsman, Gary M; Morris, Brent J; Unger, R Zackary; Laughlin, Mitzi S; Elkousy, Hussein A; Edwards, T Bradley

    2015-03-04

    The purpose of this study was to determine characteristics and trends in published shoulder research over the last decade in a leading orthopaedic journal. We examined all clinical shoulder articles published in The Journal of Bone & Joint Surgery from 2004 to 2014. The number of citations, authorship, academic degrees of the authors, country and institution of origin, topic, level of evidence, positive or nonpositive outcome, and inclusion of validated patient-reported outcome measures were assessed for each article. Shoulder articles that included an author with an advanced research degree (MD [Doctor of Medicine] with a PhD [Doctor of Philosophy] or other advanced degree) increased during the study period (p = 0.047). Level-I, II, and III studies were more likely to have an author with an advanced research degree, and Level-IV studies were more likely to have MDs only (p = 0.03). Overall, there was great variability of outcome measures, with at least thirty-nine different validated or nonvalidated outcome measures reported. Over the last decade, there was an improvement in the level of evidence of shoulder articles published in The Journal of Bone & Joint Surgery that corresponds with recent emphasis on evidence-based medicine. A consensus is needed in shoulder research for more consistent application of validated patient-reported outcome measurement tools. Copyright © 2015 by The Journal of Bone and Joint Surgery, Incorporated.

  17. Near-field non-radial motion generation from underground chemical explosions in jointed granite

    DOE PAGES

    Vorobiev, Oleg; Ezzedine, Souheil; Hurley, Ryan

    2017-09-22

    Here, this paper describes analysis of non-radial ground motion generated by chemical explosions in a jointed rock formation during the Source Physics Experiment (SPE). Such motion makes it difficult to discriminate between various subsurface events such as explosions, implosions (i.e. mine collapse) and earthquakes. We apply 3-D numerical simulations to understand experimental data collected during the SPEs. The joints are modelled explicitly as compliant thin inclusions embedded into the rock mass. Mechanical properties of the rock and the joints as well as the joint spacing and orientation are inferred from experimental test data, and geophysical and geological characterization of themore » SPE site which is dominantly Climax Stock granitic outcrop. The role of various factors characterizing the joints such as joint spacing, frictional properties, orientation and persistence in generation of non-radial motion is addressed. The joints in granite at the SPE site are oriented in nearly orthogonal directions with two vertical sets dipping at 70–80 degrees with the same strike angle, one vertical set almost orthogonal to the first two and one shallow angle joint set dipping 15 degrees. In this study we establish the relationship between the joint orientation and azimuthal variations in the polarity of the observed shear motion. The majority of the shear motion is generated due to the effects of non-elastic sliding on the joints near the source, where the wave can create significant shear stress to overcome the cohesive forces at the joints. Near the surface the joints are less confined and are subject to sliding when the pressure waves are reflected. In the far field, where the cohesive forces on the joints cannot be overcome, additional shear motion can be generated due to elastic anisotropy of the rock mass given by preferred spatial orientations of compliant joints.« less

  18. Near-field non-radial motion generation from underground chemical explosions in jointed granite

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Vorobiev, Oleg; Ezzedine, Souheil; Hurley, Ryan

    Here, this paper describes analysis of non-radial ground motion generated by chemical explosions in a jointed rock formation during the Source Physics Experiment (SPE). Such motion makes it difficult to discriminate between various subsurface events such as explosions, implosions (i.e. mine collapse) and earthquakes. We apply 3-D numerical simulations to understand experimental data collected during the SPEs. The joints are modelled explicitly as compliant thin inclusions embedded into the rock mass. Mechanical properties of the rock and the joints as well as the joint spacing and orientation are inferred from experimental test data, and geophysical and geological characterization of themore » SPE site which is dominantly Climax Stock granitic outcrop. The role of various factors characterizing the joints such as joint spacing, frictional properties, orientation and persistence in generation of non-radial motion is addressed. The joints in granite at the SPE site are oriented in nearly orthogonal directions with two vertical sets dipping at 70–80 degrees with the same strike angle, one vertical set almost orthogonal to the first two and one shallow angle joint set dipping 15 degrees. In this study we establish the relationship between the joint orientation and azimuthal variations in the polarity of the observed shear motion. The majority of the shear motion is generated due to the effects of non-elastic sliding on the joints near the source, where the wave can create significant shear stress to overcome the cohesive forces at the joints. Near the surface the joints are less confined and are subject to sliding when the pressure waves are reflected. In the far field, where the cohesive forces on the joints cannot be overcome, additional shear motion can be generated due to elastic anisotropy of the rock mass given by preferred spatial orientations of compliant joints.« less

  19. Customizing Extensor Reconstruction in Vascularized Toe Joint Transfers to Finger Proximal Interphalangeal Joints: A Strategic Approach for Correcting Extensor Lag.

    PubMed

    Loh, Charles Yuen Yung; Hsu, Chung-Chen; Lin, Cheng-Hung; Chen, Shih-Heng; Lien, Shwu-Huei; Lin, Chih-Hung; Wei, Fu-Chan; Lin, Yu-Te

    2017-04-01

    Vascularized toe proximal interphalangeal joint transfer allows the restoration of damaged joints. However, extensor lag and poor arc of motion have been reported. The authors present their outcomes of treatment according to a novel reconstructive algorithm that addresses extensor lag and allows for consistent results postoperatively. Vascularized toe joint transfers were performed in a consecutive series of 26 digits in 25 patients. The average age was 30.5 years, with 14 right and 12 left hands. Reconstructed digits included eight index, 10 middle, and eight ring fingers. Simultaneous extensor reconstructions were performed and eight were centralization of lateral bands, five were direct extensor digitorum longus-to-extensor digitorum communis repairs, and 13 were central slip reconstructions. The average length of follow-up was 16.7 months. The average extension lag was 17.9 degrees. The arc of motion was 57.7 degrees (81.7 percent functional use of pretransfer toe proximal interphalangeal joint arc of motion). There was no significant difference in the reconstructed proximal interphalangeal joint arc of motion for the handedness (p = 0.23), recipient digits (p = 0.37), or surgical experience in vascularized toe joint transfer (p = 0.25). The outcomes of different techniques of extensor mechanism reconstruction were similar in terms of extensor lag, arc of motion, and reconstructed finger arc of motion compared with the pretransfer toe proximal interphalangeal joint arc of motion. With this treatment algorithm, consistent outcomes can be produced with minimal extensor lag and maximum use of potential toe proximal interphalangeal joint arc of motion. Therapeutic, IV.

  20. New design for a rotatory joint actuator made with shape memory alloy contractile wire

    NASA Astrophysics Data System (ADS)

    Wang, Guoping; Shahinpoor, Mohsen

    1996-05-01

    A design approach for a rotatory joint actuator using a contractile shape memory alloy (SMA) wire is presented and an example design is followed. In this example, the output torque of the actuator is 18 Newton-meters, and its angular range is 30 degrees. Compared with a SMA spring type actuating component, a SMA wire type actuating component uses less SMA material and uses less electrical energy when it is electrically powered. On the other hand, a SMA wire type actuating component must have a large SMA wire length to produce a required amount of angular rotation of the joint. When pulleys are used to arrange a lengthy SMA wire in a small space, the friction between pulleys and pins is introduced and the performance of the joint actuator is degenerated to some degree. The investigated joint actuator provides a good chance for developing powered orthoses with SMA actuators for disabled individuals. It can relieve the weight concern with hydraulic and motor-powered orthoses and the safety concern with motor-powered orthoses. When electrically powered, a SMA actuator has the disadvantage of low energy efficiency.

  1. Research regarding the influence of driving-wires length change on positioning precision of a robotic arm

    NASA Astrophysics Data System (ADS)

    Ciofu, C.; Stan, G.

    2016-08-01

    The paper emphasise positioning precision of an elephant's trunk robotic arm which has joints driven by wires with variable length while operating The considered 5 degrees of freedom robotic arm has a particular structure of joint that makes possible inner actuation with wire-driven mechanism. We analyse solely the length change of wires as a consequence due inner winding and unwinding on joints for certain values of rotational angles. Variations in wires length entail joint angular displacements. We analyse positioning precision by taking into consideration equations from inverse kinematics of the elephant's trunk robotic arm. The angular displacements of joints are considered into computational method after partial derivation of positioning equations. We obtain variations of wires length at about tenths of micrometers. These variations employ angular displacements which are about minutes of sexagesimal degree and, thus, define positioning precision of elephant's trunk robotic arms. The analytical method is used for determining aftermath design structure of an elephant's trunk robotic arm with inner actuation through wires on positioning precision. Thus, designers could take suitable decisions on accuracy specifications limits of the robotic arm.

  2. Kinematic equations for control of the redundant eight-degree-of-freedom advanced research manipulator 2

    NASA Technical Reports Server (NTRS)

    Williams, Robert L., II

    1992-01-01

    The forward position and velocity kinematics for the redundant eight-degree-of-freedom Advanced Research Manipulator 2 (ARM2) are presented. Inverse position and velocity kinematic solutions are also presented. The approach in this paper is to specify two of the unknowns and solve for the remaining six unknowns. Two unknowns can be specified with two restrictions. First, the elbow joint angle and rate cannot be specified because they are known from the end-effector position and velocity. Second, one unknown must be specified from the four-jointed wrist, and the second from joints that translate the wrist, elbow joint excluded. There are eight solutions to the inverse position problem. The inverse velocity solution is unique, assuming the Jacobian matrix is not singular. A discussion of singularities is based on specifying two joint rates and analyzing the reduced Jacobian matrix. When this matrix is singular, the generalized inverse may be used as an alternate solution. Computer simulations were developed to verify the equations. Examples demonstrate agreement between forward and inverse solutions.

  3. Joint with application in electrochemical devices

    DOEpatents

    Weil, K Scott [Richland, WA; Hardy, John S [Richland, WA

    2010-09-14

    A joint for use in electrochemical devices, such as solid oxide fuel cells (SOFCs), oxygen separators, and hydrogen separators, that will maintain a hermetic seal at operating temperatures of greater than 600.degree. C., despite repeated thermal cycling excess of 600.degree. C. in a hostile operating environment where one side of the joint is continuously exposed to an oxidizing atmosphere and the other side is continuously exposed to a wet reducing gas. The joint is formed of a metal part, a ceramic part, and a flexible gasket. The flexible gasket is metal, but is thinner and more flexible than the metal part. As the joint is heated and cooled, the flexible gasket is configured to flex in response to changes in the relative size of the metal part and the ceramic part brought about by differences in the coefficient of thermal expansion of the metal part and the ceramic part, such that substantially all of the tension created by the differences in the expansion and contraction of the ceramic and metal parts is absorbed and dissipated by flexing the flexible gasket.

  4. Experimental joint immobilization in guinea pigs. Effects on the knee joint

    NASA Technical Reports Server (NTRS)

    Marcondesdesouza, J. P.; Machado, F. F.; Sesso, A.; Valeri, V.

    1980-01-01

    In young and adult guinea pigs, the aftermath experimentally induced by the immobilization of the knee joint in hyperextended forced position was studied. Joint immobilization which varied from one to nine weeks was attained by plaster. Eighty knee joints were examined macro and microscopically. Findings included: (1) muscular hypotrophy and joint stiffness in all animals, directly proportional to the length of immobilization; (2) haemoarthrosis in the first week; (3) intra-articular fibrous tissue proliferation ending up with fibrous ankylosis; (4) hyaline articular cartilage erosions; (5) various degrees of destructive menisci changes. A tentative explanation of the fibrous tissue proliferation and of the cartilage changes is offered.

  5. Joint DoD versus Navy Specific Lead Generation Advertising: Comparison of Conversion Rates to Quality Enlistments and Marginal Costs.

    DTIC Science & Technology

    1984-09-01

    7D-Rt46 982 JOINT DOD VERSUS NAVY SPECIFIC LEAD GENERATION j/j ADVERTISING : COMPARISON OF..(U) J B FUGUR SCHOOL OF N BUSINESS DURHAM NC R C MOREY...REPORT I PEPIO0 COV9cO JOINT DOD VERSUS NAVY SPECIFIC LEAD GENERATION Technical Report ADVERTISING : Comparison of Conversion Rates to (0 Quality...block number) . Upper-Mental, High School Degree, enlistment contracts, national leads, Z Joint DOD advertising , Service Specific Advertising , conversion

  6. Properties of welded joints in laser welding of aeronautic aluminum-lithium alloys

    NASA Astrophysics Data System (ADS)

    Malikov, A. G.; Orishich, A. M.

    2017-01-01

    The work presents the experimental investigation of the laser welding of the aluminum-lithium alloys (system Al-Mg-Li) and aluminum alloy (system Al-Cu-Li) doped with Sc. The influence of the nano-structuring of the surface layer welded joint by the cold plastic deformation method on the strength properties of the welded joint is determined. It is founded that, regarding the deformation degree over the thickness, the varying value of the welded joint strength is different for these aluminum alloys.

  7. Risk management in international manned space program operations.

    PubMed

    Seastrom, J W; Peercy, R L; Johnson, G W; Sotnikov, B J; Brukhanov, N

    2004-02-01

    New, innovative joint safety policies and requirements were developed in support of the Shuttle/Mir program, which is the first phase of the International Space Station program. This work has resulted in a joint multinational analysis culminating in joint certification for mission readiness. For these planning and development efforts, each nation's risk programs and individual safety practices had to be integrated into a comprehensive and compatible system that reflects the joint nature of the endeavor. This paper highlights the major incremental steps involved in planning and program integration during development of the Shuttle/Mir program. It traces the transition from early development to operational status and highlights the valuable lessons learned that apply to the International Space Station program (Phase 2). Also examined are external and extraneous factors that affected mission operations and the corresponding solutions to ensure safe and effective Shuttle/Mir missions. c2003 Published by Elsevier Ltd.

  8. Joint Task Force on Undergraduate Physics Programs (J-TUPP): What skills and knowledge are needed for a diverse set of careers and what's the basis for these recommendations?

    NASA Astrophysics Data System (ADS)

    Woolf, Lawrence

    2016-03-01

    A wide variety of reports have been issued recently concerning the skills, knowledge, and attitudes needed by employees to be successful. This talk will review findings from reports from the major science and engineering disciplines, from surveys of employers, and from interviews with recent undergraduate physics graduates. Also to be discussed is the correlation between these findings and the detailed J-TUPP recommendations for the skills and knowledge needed by the next generation of undergraduate physics degree holders to be prepared for a diverse set of careers.

  9. [Analysis on the long-term effects of modified double endobutton technique in the treatment of Tossy type III acromioclavicular joint dislocations].

    PubMed

    Yan, Rui-Jian; Lu, Jian-Wei; Zhang, Chun

    2014-01-01

    To investigate the long-term clinical effects of modified double Endobutton technique for the treatment of acromioclavicular joint dislocations of Tossy type III. A retrospective study was done in 42 patients with acromioclavicular joint dislocations of Tossy type III treated with modified double Endobutton technique from December 2008 to December 2010. There were 24 males and 18 females, ranging in age from 21 to 56 years old (averaged, 32.5 years old). All the patients were treated with open reduction, coracoclavicular ligament reconstruction using double Endobutton technique, and repair of acromioclavicular ligament. The Karlsson system was used to evaluate therapeutic effects. The distance from coracoid to clavicle was measured to evaluate reduction loss. All the patients were followed up, and the duration ranged from 2.0 to 3.2 years (averaged,2.4 years). According to Karlsson system, 32 patients got an A degree and 10 patients got a B degree at three months post-operatively; 26 patients got an A degree and 16 patients got a B degree at the latest follow-up; 6 patients got an A degree at 3 months after operation lowered to B degree at the latest follow-up. The coracoid-clavicle distance increased from (26.91 +/- 0.91) mm at 3 months after operation to (27.41 +/- 1.10) mm at the latest follow-up. Te patients treated with over-reduction during operation or with heavy physical labour work after operation had obvious widened coracoid-clavicle distance. Bone absorption was found around the plate in most cases, mainly in the clavicular side. Treatment for acromioclavicular joint dislocations of Tossy type III with modified double Endobutton technique has satisfactory early clinical results. But with time passing, loss of reduction and bone absorption around the plate could be observed, and clinical outcomes of some cases downgrade during the long-term follow-up.

  10. [Research in the PhD Program led by János Fehér between 1993 and 2010 at the Biochemical Research Laboratory, 2nd Department of Medicine, Semmelweis University].

    PubMed

    Blázovics, Anna

    2010-11-21

    Author wish to express gratitude to late professor János Fehér for the invitation to participate in "Free Radical and Immunological References of Hepatology" PhD program in 1993 and for providing opportunity to establish a laboratory at the 2nd Department of Medicine, Semmelweis University. He established a joint medical and biological research that is continuing unbrokenly. In this research group, between 1993 and 2010, eleven Ph.D. students received their scientific degrees and two candidate dissertations were prepared. Three students are working in this very exciting field even today. Author would like to salute before János Fehér's remembrance by giving a list of results of topics under her leadership.

  11. Reliability of the standard goniometry and diagrammatic recording of finger joint angles: a comparative study with healthy subjects and non-professional raters.

    PubMed

    Macionis, Valdas

    2013-01-09

    Diagrammatic recording of finger joint angles by using two criss-crossed paper strips can be a quick substitute to the standard goniometry. As a preliminary step toward clinical validation of the diagrammatic technique, the current study employed healthy subjects and non-professional raters to explore whether reliability estimates of the diagrammatic goniometry are comparable with those of the standard procedure. The study included two procedurally different parts, which were replicated by assigning 24 medical students to act interchangeably as 12 subjects and 12 raters. A larger component of the study was designed to compare goniometers side-by-side in measurement of finger joint angles varying from subject to subject. In the rest of the study, the instruments were compared by parallel evaluations of joint angles similar for all subjects in a situation of simulated change of joint range of motion over time. The subjects used special guides to position the joints of their left ring finger at varying angles of flexion and extension. The obtained diagrams of joint angles were converted to numerical values by computerized measurements. The statistical approaches included calculation of appropriate intraclass correlation coefficients, standard errors of measurements, proportions of measurement differences of 5 or less degrees, and significant differences between paired observations. Reliability estimates were similar for both goniometers. Intra-rater and inter-rater intraclass correlation coefficients ranged from 0.69 to 0.93. The corresponding standard errors of measurements ranged from 2.4 to 4.9 degrees. Repeated measurements of a considerable number of raters fell within clinically non-meaningful 5 degrees of each other in proportions comparable with a criterion value of 0.95. Data collected with both instruments could be similarly interpreted in a simulated situation of change of joint range of motion over time. The paper goniometer and the standard goniometer can be used interchangeably by non-professional raters for evaluation of normal finger joints. The obtained results warrant further research to assess clinical performance of the paper strip technique.

  12. Reliability of the standard goniometry and diagrammatic recording of finger joint angles: a comparative study with healthy subjects and non-professional raters

    PubMed Central

    2013-01-01

    Background Diagrammatic recording of finger joint angles by using two criss-crossed paper strips can be a quick substitute to the standard goniometry. As a preliminary step toward clinical validation of the diagrammatic technique, the current study employed healthy subjects and non-professional raters to explore whether reliability estimates of the diagrammatic goniometry are comparable with those of the standard procedure. Methods The study included two procedurally different parts, which were replicated by assigning 24 medical students to act interchangeably as 12 subjects and 12 raters. A larger component of the study was designed to compare goniometers side-by-side in measurement of finger joint angles varying from subject to subject. In the rest of the study, the instruments were compared by parallel evaluations of joint angles similar for all subjects in a situation of simulated change of joint range of motion over time. The subjects used special guides to position the joints of their left ring finger at varying angles of flexion and extension. The obtained diagrams of joint angles were converted to numerical values by computerized measurements. The statistical approaches included calculation of appropriate intraclass correlation coefficients, standard errors of measurements, proportions of measurement differences of 5 or less degrees, and significant differences between paired observations. Results Reliability estimates were similar for both goniometers. Intra-rater and inter-rater intraclass correlation coefficients ranged from 0.69 to 0.93. The corresponding standard errors of measurements ranged from 2.4 to 4.9 degrees. Repeated measurements of a considerable number of raters fell within clinically non-meaningful 5 degrees of each other in proportions comparable with a criterion value of 0.95. Data collected with both instruments could be similarly interpreted in a simulated situation of change of joint range of motion over time. Conclusions The paper goniometer and the standard goniometer can be used interchangeably by non-professional raters for evaluation of normal finger joints. The obtained results warrant further research to assess clinical performance of the paper strip technique. PMID:23302419

  13. Pre-irradiation testing of actively cooled Be-Cu divertor modules

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Linke, J.; Duwe, R.; Kuehnlein, W.

    1995-09-01

    A set of neutron irradiation tests is prepared on different plasma facing materials (PFM) candidates and miniaturized components for ITER. Beside beryllium the irradiation program which will be performed in the High Flux Reactor (HFR) in Petten, includes different carbon fiber composites (CFQ) and tungsten alloys. The target values for the neutron irradiation will be 0.5 dpa at temperatures of 350{degrees}C and 700{degrees}C, resp.. The post irradiation examination (PIE) will cover a wide range of mechanical tests; in addition the degradation of thermal conductivity will be investigated. To determine the high heat flux (HHF) performance of actively cooled divertor modules,more » electron beam tests which simulate the expected heat loads during the operation of ITER, are scheduled in the hot cell electron beam facility JUDITH. These tests on a selection of different actively cooled beryllium-copper and CFC-copper divertor modules are performed before and after neutron irradiation; the pre-irradiation testing is an essential part of the program to quantify the zero-fluence high heat flux performance and to detect defects in the modules, in particular in the brazed joints.« less

  14. [Restricted motion after total knee arthroplasty].

    PubMed

    Kucera, T; Urban, K; Karpas, K; Sponer, P

    2007-10-01

    The aim of the study was to ascertain what proportion of patients undergoing total knee arthroplasty (TKA) complain of restricted knee joint motion, and to investigate options for improvement of this situation. Our evaluation included a group of 796 patients treated with TKA at our department in the period from January 1, 1990, to December 31, 2004. In all cases, a condylar implant with preservation of the posterior cruciate ligaments was used. In addition to medical history, the range of motion, knee joint malalignment and radiological findings were assessed before surgery. After THA, the type of implant and complications, if any, were recorded, and improvement in joint motion was followed up. Based on the results of Kim et al., flexion contracture equal to or higher than 15 degrees and/or flexion less than 75 degrees were made the criteria of stiffness after THA. Patients with restricted THA motion who had aseptic or septic implant loosening were not included. Of the 796 evaluated patients, 32 (4.14 %) showed restricted motion after total knee arthroplasty, as assessed by the established criteria. In 16 patients, stiffness defined by these criteria had existed before surgery, and three patients showed an excessive production of adhesions and heterotopic ossifications. In three patients, the implantation procedure resulted in an elevated level of the original joint line and subsequent development of patella infera and increased tension of the posterior cruciate ligament. Four patients declined physical therapy and, in six, the main cause of stiffness could not be found. Seventeen patients did not require surgical therapy for restricted motion; TKA provided significant pain relief and they considered the range of motion achieved to be sufficient. One patient underwent redress 3 months after surgery, but with no success. Repeated releases of adhesions, replacement of a polyethylene liner and revision surgery of the extensor knee structures were performed in 15 patients. In these, the average value of knee flexion increased by 17 degrees only and, in the patients suffering from excessive adhesion production, this value remained almost unchanged. Revision TKA was carried out in four patients, in whom knee joint flexion increased on average by 35 degrees to achieve an average flexion of 83 degrees. Restricted motion after TKA has been reported to range from 1.3 % to 12.0 %, but consistent criteria have not been set up. In our study it was 4.14 %. In agreement with the literature data, one of the reasons was pre-operative restricted motion, which was recorded in 16 of 32 patients. Similarly, also in our patients, biological predisposition to excessive production of fibrocartilage associated with adhesions in all knee joint compartments was the major therapeutic problem. Intra-operative fractures, ligament tears requiring post-operative fixation and unremoved dorsal osteophytes lead to the restriction of knee joint motion. By inadequate resection of articular surface, the original joint line may be at a higher level; this results in an increased tension of the posterior cruciate ligament and patella infera development, both influencing knee flexion. In our study, three patients were affected. Knee joint stiffness can also develop in patients declining physical therapy or in whom this is not correctly performed, often for insufficient analgesia. In contrast to the data reported in the literature, 17 of 32 patients in this study had no need for surgical treatment of restricted knee joint motion. Redress under general anesthesia was not effective. For markedly restricted motion of the knee joint, reimplantation can be recommended or, in less severe cases, an intervention on adjacent soft tissues. Restricted motion of the knee joint after TKA is difficult to treat and, therefore, prevention is recommended. This should include thorough conservative treatment of gonarthrosis, early indication for surgery, prevention of elevation in the joint line and consistent rehabilitation with appropriate analgesia. For severe stiffness of the knee joint, as evaluated by the criteria of Kim et al., revision arthroplasty can be recommended.

  15. Delayed latency of peroneal reflex to sudden inversion with ankle taping or bracing.

    PubMed

    Shima, N; Maeda, A; Hirohashi, K

    2005-01-01

    The purpose of the present study was to examine the effects of ankle taping and bracing based on the peroneal reflex in the hypermobile and normal ankle joints with and without history of ankle injury. Thirty-six ankle joints of 18 collegiate American football athletes with and without previous history of injury were studied. The angle of talar tilt (TT) was measured by stress radiograph for classifying normal (TT5 degrees ) ankles. They were tested with taping, bracing, and without any supports as a control. The latency of peroneus longus muscle was measured by a sudden inversion of 25 degrees using surface EMG signals. The results of the present study show no significant three-way Group (hypermobile or normal ankles) by History (previously injured or uninjured ankles) by Condition (control, taping, or bracing) interaction, while Condition main effect was significant (p<0.05). There were significant differences between control (80.8 ms) and taping (83.8 ms, p<0.01), between control and bracing (83.0 ms, p<0.05), but not between taping and bracing (p>0.05). In conclusion, ankle taping and bracing delayed the peroneal reflex latency not only for hypermobile ankles and/or injured ankle joints but also for intact ankle joints.

  16. Proprioception and throwing accuracy in the dominant shoulder after cryotherapy.

    PubMed

    Wassinger, Craig A; Myers, Joseph B; Gatti, Joseph M; Conley, Kevin M; Lephart, Scott M

    2007-01-01

    Application of cryotherapy modalities is common after acute shoulder injury and as part of rehabilitation. During athletic events, athletes may return to play after this treatment. The effects of cryotherapy on dominant shoulder proprioception have been assessed, yet the effects on throwing performance are unknown. To determine the effects of a cryotherapy application on shoulder proprioception and throwing accuracy. Single-group, pretest-posttest control session design. University-based biomechanics laboratory. Healthy college-aged subjects (n = 22). Twenty-minute ice pack application to the dominant shoulder. Active joint position replication, path of joint motion replication, and the Functional Throwing Performance Index. Subjects demonstrated significant increases in deviation for path of joint motion replication when moving from 90 degrees of abduction with 90 degrees of external rotation to 20 degrees of flexion with neutral shoulder rotation after ice pack application. Also, subjects exhibited a decrease in Functional Throwing Performance Index after cryotherapy application. No differences were found in subjects for active joint position replication after cryotherapy application. Proprioception and throwing accuracy were decreased after ice pack application to the shoulder. It is important that clinicians understand the deficits that occur after cryotherapy, as this modality is commonly used following acute injury and during rehabilitation. This information should also be considered when attempting to return an athlete to play after treatment.

  17. Station corrections for the Katmai Region Seismic Network

    USGS Publications Warehouse

    Searcy, Cheryl K.

    2003-01-01

    Most procedures for routinely locating earthquake hypocenters within a local network are constrained to using laterally homogeneous velocity models to represent the Earth's crustal velocity structure. As a result, earthquake location errors may arise due to actual lateral variations in the Earth's velocity structure. Station corrections can be used to compensate for heterogeneous velocity structure near individual stations (Douglas, 1967; Pujol, 1988). The HYPOELLIPSE program (Lahr, 1999) used by the Alaska Volcano Observatory (AVO) to locate earthquakes in Cook Inlet and the Aleutian Islands is a robust and efficient program that uses one-dimensional velocity models to determine hypocenters of local and regional earthquakes. This program does have the capability of utilizing station corrections within it's earthquake location proceedure. The velocity structures of Cook Inlet and Aleutian volcanoes very likely contain laterally varying heterogeneities. For this reason, the accuracy of earthquake locations in these areas will benefit from the determination and addition of station corrections. In this study, I determine corrections for each station in the Katmai region. The Katmai region is defined to lie between latitudes 57.5 degrees North and 59.00 degrees north and longitudes -154.00 and -156.00 (see Figure 1) and includes Mount Katmai, Novarupta, Mount Martin, Mount Mageik, Snowy Mountain, Mount Trident, and Mount Griggs volcanoes. Station corrections were determined using the computer program VELEST (Kissling, 1994). VELEST inverts arrival time data for one-dimensional velocity models and station corrections using a joint hypocenter determination technique. VELEST can also be used to locate single events.

  18. Kinematic synthesis of bevel-gear-type robotic wrist mechanisms

    NASA Astrophysics Data System (ADS)

    Lin, Chen-Chou

    Bevel-gear-type robotic wrist mechanisms are commonly used in industry. The reasons for their popularity are that they are compact, light-weight, and relatively inexpensive. However, there are singularities in their workspace, which substantially degrade their manipulative performance. The objective of this research is to develop an atlas of three-degree-of-freedom bevel-gear-type wrist mechanisms, and through dimensional synthesis to improve their kinematic performance. The dissertation contains two major parts: the first is structural analysis and synthesis, the other is kinematic analysis and dimensional synthesis. To synthesize the kinematic structures of bevel-gear-type wrist mechanisms, the kinematic structures are separated from their functional considerations. All kinematic structures which satisfy the mobility condition are enumerated in an unbiased, systematic manner. Then the bevel-gear-type wrist mechanisms are identified by applying the functional requirements. Structural analysis shows that a three-degree-of-freedom wrist mechanism usually consists of non-fractionated, two degree-of-freedom epicyclic gear train jointed with the base link. Therefore, the structural synthesis can be simplified into a problem of examining the atlas of non-fractionated, two-degree-of-freedom epicyclic gear trains. The resulting bevel-gear-type wrist mechanism has been categorized and evaluated. It is shown that three-degree-of-freedom, four-jointed wrist mechanisms are promising for further improving the kinematic performance. It is found that a spherical planetary gear train is necessarily imbedded in a three-degree-of-freedom, four-jointed wrist mechanism. Therefore, to study the workspace and singularity problems of three-degree-of-freedom four-jointed spherical wrist mechanisms, we have to study the trajectories of spherical planetary gear trains. The parametric equations of the trajectories and some useful geometric properties for the analysis and synthesis of workplace are derived. The workspace boundary equations can be derived via both geometric consideration and Jacobian analysis. The workspace is divided by inner and outer boundaries into regions of accessibility of zero, two, and four. The design criteria of full workspace and a maximum four-root region are established.

  19. Shape of wear particles found in human knee joints and their relationship to osteoarthritis.

    PubMed

    Kuster, M S; Podsiadlo, P; Stachowiak, G W

    1998-09-01

    To analyse and compare the shape of wear particles found in healthy and osteoarthritic human knee joints for monitoring the progress of osteoarthritis, the long-term prognosis and to evaluate therapeutic regimens. Joint particles from seven patients with normal cartilage in all compartments of the knee joint, 12 patients with fibrillation of less than half the cartilage thickness (grade 1), seven patients with fibrillation of more than half the cartilage thickness (grade 2) and four patients with erosions down to bone (grade 3) were analysed. A total of 565 particles were extracted from synovial fluid samples by ferrography and analysed in a scanning electron microscope. A number of numerical descriptors, i.e. boundary fractal dimension, shape factor, convexity and elongation, were calculated for each particle image and correlated to the degree of osteoarthritis using non-parametric tests. Experiments demonstrated that there were significant differences between the numerical descriptors calculated for wear particles from healthy and osteoarthritic knee joints (P < 0.01), suggesting that the particle shape can be used as an indicator of the joint condition. In particular, the fractal dimension of the particle boundary was shown to correlate directly with the degree of osteoarthritis. Numerical analysis of the shape of wear particles found in human knee joints may provide a reliable means for the assessment of cartilage repair after surgical or conservative treatment of osteoarthritis.

  20. Extra-articular osteotomy for malunited unicondylar fractures of the proximal phalanx.

    PubMed

    Harness, Neil G; Chen, Alvin; Jupiter, Jesse B

    2005-05-01

    To evaluate an extra-articular osteotomy rather than an intra-articular osteotomy in the treatment of malunited unicondylar fractures of the proximal phalanx. An extra-articular osteotomy was used to correct the deformity resulting from a malunion of a unicondylar fracture of the proximal phalanx in 5 patients. A closing wedge osteotomy that was stabilized with tension band fixation accomplished realignment of the joint. Each patient was evaluated at a minimum of 1 year after surgery for radiographic healing, correction of angulation, digital motion, postoperative complications, current level of pain with motion, and overall satisfaction with the procedure. All of the osteotomies healed by 10 to 12 weeks after surgery with an average angular correction from 25 degrees to 1 degrees . The average proximal interphalangeal joint motion improved to 86 degrees from the preoperative average of 40 degrees , whereas the average total digital motion improved from 154 degrees before surgery to 204 degrees at follow-up evaluation. This method of extra-articular osteotomy for malunited unicondylar fractures of the proximal phalanx is highly reproducible, avoids the risks of intra-articular surgery, and leads to a predictable outcome.

  1. A Model for a Single Unmanned Aircraft Systems (UAS) Program Office Managing Joint ISR Capabilities

    DTIC Science & Technology

    2017-10-01

    reduction in manning from the multiple program office structure to the new single program management model. Additional information regarding this...OFFICE MANAGING JOINT ISR CAPABILITIES by Angela E. Burris A Research Report Submitted to the Faculty In Partial Fulfillment of...research paper is to answer how a single management office could provide greater agility for unmanned aircraft systems (UAS); supporting Joint concepts

  2. Mechanistic analysis of time-dependent failure of oxynitride glass-joined silicon nitride below 1000 degree C

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    O'Brien, M.H.; Coon, D.M.

    Time-dependent failure at elevated temperatures currently governs the service life of oxynitride glass-joined silicon nitride. Creep, devitrification, stress- aided oxidation-controlled slow crack growth, and viscous cabitation-controlled failure are examined as possible controlling mechanisms. Creep deformation failure is observed above 1000{degrees}C. Fractographic evidence indicates cavity formation and growth below 1000{degrees}C. Auger electron spectroscopy verified that the oxidation rate of the joining glass is governed by the oxygen supply rate. Time-to-failure data and those predicted using the Tsai and Raj, and Raj and Dang viscous cavitation models. It is concluded that viscous relaxation and isolated cavity growth control the rate of failuremore » in oxynitride glass-filled silicon nitride joints below 1000{degrees}C. Several possible methods are also proposed for increasing the service lives of these joints.« less

  3. Use of the dual force system to correct chronic knee deformities due to severe haemophilia.

    PubMed

    Kale, J S; Ghosh, K; Mohanty, D; Pathare, A V; Jijina, F

    2000-05-01

    In this study, the use of the dual force system to correct recent or relatively longstanding knee deformities in ten patients is described. (Nine of the patients had severe haemophilia and one had severe von Willebrand's disease.) The mean duration of deformity in these patients was 10 months. The mean range of movement at the affected knee joints increased from 50 degrees at pre-intervention to 110 degrees following 6 weeks of application of the dual force system. In nine of ten patients (90%) the residual flexion deformity ranged from 0 degrees to 10 degrees. The dual force system offers an easily affordable and effective means of correcting a flexion deformity of the knee joint in severely affected haemophilia and allied disorders. More extensive use of this technique in different centres is required to determine its place in the day-to-day management of such patients.

  4. [Treatment of fractures of proximal phalanx of fingers by Eiffel Tower percutaneous pinning method. A review of 45 cases].

    PubMed

    Chbani, B; Amar, M F; Loudyi, D; Boutayeb, F

    2010-04-01

    The authors report in the treatment of fractures of the proximal phalanx of the fingers, the use of Eiffel Tower pinning, a relatively simple method, fast and stable, associated to a protection and early rehabilitation. The objective of this method is to offer to the patient a pollici-digital grip. Our study is a retrospective study of 45 patients treated for fractures of the proximal phalanx of the fingers by percutaneous pinning according to Eiffel Tower method. We detail this simple and economic technique and examine the functional and radiological results of this series of patients. The amplitude of the active total motion of the proximal interphalangeal joint is on average 94.16 degrees (78.5 % of the normal active mobility of the proximal interphalangeal joint), and the amplitude of the active total motion of the metacarpo-phalangeal joint is on average 90.05 degrees (75 % of the normal active mobility of the metacarpo-phalangeal joint). Copyright 2010 Elsevier Masson SAS. All rights reserved.

  5. Low-Friction, High-Stiffness Joint for Uniaxial Load Cell

    NASA Technical Reports Server (NTRS)

    Lewis, James L.; Le, Thang; Carroll, Monty B.

    2007-01-01

    A universal-joint assembly has been devised for transferring axial tension or compression to a load cell. To maximize measurement accuracy, the assembly is required to minimize any moments and non-axial forces on the load cell and to exhibit little or no hysteresis. The requirement to minimize hysteresis translates to a requirement to maximize axial stiffness (including minimizing backlash) and a simultaneous requirement to minimize friction. In practice, these are competing requirements, encountered repeatedly in efforts to design universal joints. Often, universal-joint designs represent compromises between these requirements. The improved universal-joint assembly contains two universal joints, each containing two adjustable pairs of angular-contact ball bearings. One might be tempted to ask why one could not use simple ball-and-socket joints rather than something as complex as universal joints containing adjustable pairs of angularcontact ball bearings. The answer is that ball-and-socket joints do not offer sufficient latitude to trade stiffness versus friction: the inevitable result of an attempt to make such a trade in a ball-and-socket joint is either too much backlash or too much friction. The universal joints are located at opposite ends of an axial subassembly that contains the load cell. The axial subassembly includes an axial shaft, an axial housing, and a fifth adjustable pair of angular-contact ball bearings that allows rotation of the axial housing relative to the shaft. The preload on each pair of angular-contact ball bearings can be adjusted to obtain the required stiffness with minimal friction, tailored for a specific application. The universal joint at each end affords two degrees of freedom, allowing only axial force to reach the load cell regardless of application of moments and non-axial forces. The rotational joint on the axial subassembly affords a fifth degree of freedom, preventing application of a torsion load to the load cell.

  6. Masters International Program in Natural Hazards: a joint program between Michigan Tech and US Peace Corps begins move to work in Indonesia

    NASA Astrophysics Data System (ADS)

    Rose, W. I.; Carn, S. A.; Waite, G. P.; Gierke, J. S.; Wellik, J. J.

    2011-12-01

    We are in the seventh year of developing a unique graduate degree program in which each student serves in the U.S. Peace Corps for two years while conducting his/her field research. Our program allows candidates to work on natural hazard mitigation projects in a country where natural hazards are important parts of life. For US students, living abroad provides a vital broadening experience and the Peace Corps emphasis on social context adds cultural understanding to their hazards work. Up until now, we have mostly worked in Central America, and 33 students have enrolled in the program. The greatest focus to date has been in Volcanic Hazards, including slope stability and debris flows, and our work is fostering long-term infrastructure-building relationships with partner agencies within the 8 countries where we have worked. This year we sent a student (Jay Wellik) to a Peace Corps site in East Java, Indonesia where he will work with schools in his village and commute weekly to the Raung Observatory Post to work with CVGHM scientists on volcano seismology and public outreach projects.. We recruit 4-6 new students each year, and we hope more will soon be in Indonesia as Peace Corps expands their new program in that country. Although the Peace Corps Masters International (PCMI) students must be US citizens, we also have regular undergraduate and graduate (MS and PhD) degree students in geology, geological engineering and geophysics who come from all over the world. We are especially interested in people from partner Peace Corps countries. Annually our natural-hazards group consists of 5 faculty, 2 post-doctoral researchers, several Ph.D and traditional M.S. students, 12 PCMI students, and roughly 20 undergraduate students. Support for our program has come from NSF and we have also benefitted from a supportive cooperation with USGS VDAP. In the past two years we have built a complementary dual degree partnership with the Université Blaise-Pascal, Clermont Ferrand (France), and Universitá degli Studi di Milano - Bicocca (Italy) , which allows for a strong European connection to this international work.

  7. Whatever Happened to Joint Vision 2010

    DTIC Science & Technology

    2010-04-02

    submitted to the Faculty of the Joint Advanced Warfighting School in partial satisfaction of the requirements of a Master of Science Degree in Joint...Service efforts and evolve “jointness” beyond the dictates of the 1986 Goldwater-Nichols Act. JV2010 delineated a common set of environmental ...towards something that is fresh, new and important. In this sense, the term vision may have been the wrong term. Warren Bennis and Burt Nanus, noted

  8. Gastrocnemius tightness on joint angle and work of lower extremity during gait.

    PubMed

    You, Jia-Yuan; Lee, Hsin-Min; Luo, Hong-Ji; Leu, Chwan-Chin; Cheng, Pen-Gang; Wu, Shyi-Kuen

    2009-11-01

    Muscular tightness is a common clinical musculoskeletal disorder and is regarded as a predisposing factor for muscle injuries. In this study, a two-way mixed design ANOVA was applied to investigate the effects of the gastrocnemius tightness on the joint angle and joint work during walking. Twenty-two patients with muscular tightness of gastrocnemius muscle (<12 degrees of ankle dorsiflexion with knee extended) and 22 age- and gender-matched subjects with normal gastrocnemius flexibility (>15 degrees of ankle dorsiflexion with knee extended) participated in this study. The joint angle and work at hip, knee, and ankle joints during the stance phase were analyzed at two preset cadences of 100 steps/min and 140 steps/min. Significantly greater flexion angles at hip (P=0.025) and knee (P=0.001) were found in the tightness group at the time of maximal ankle dorsiflexion. Significantly less work generation at knee (P=0.034) and greater work absorption at ankle (P=0.024) were detected in the tightness group. The subjects with gastrocnemius tightness revealed a compensatory gait pattern, which included the changes in the joint angles and associated work productions. The potential disturbance of the knee control and strain injuries of plantar flexors might be crucial in the clinical considerations for subjects with gastrocnemius tightness.

  9. A Mating of Tarantulas

    ERIC Educational Resources Information Center

    Lewis, Robert L.

    1978-01-01

    The role of the trustee in interinstitutional cooperation is discussed. Benefits include: plant facilities sharing, joint credit unions, library exchange, joint health care facilities, joint counseling, joint apprenticeship programs, cross registration, faculty exchange joint appointments, etc. (Author/LBH)

  10. Running in ostriches (Struthio camelus): three-dimensional joint axes alignment and joint kinematics.

    PubMed

    Rubenson, Jonas; Lloyd, David G; Besier, Thor F; Heliams, Denham B; Fournier, Paul A

    2007-07-01

    Although locomotor kinematics in walking and running birds have been examined in studies exploring many biological aspects of bipedalism, these studies have been largely limited to two-dimensional analyses. Incorporating a five-segment, 17 degree-of-freedom (d.f.) kinematic model of the ostrich hind limb developed from anatomical specimens, we quantified the three-dimensional (3-D) joint axis alignment and joint kinematics during running (at approximately 3.3 m s(-1)) in the largest avian biped, the ostrich. Our analysis revealed that the majority of the segment motion during running in the ostrich occurs in flexion/extension. Importantly, however, the alignment of the average flexion/extension helical axes of the knee and ankle are rotated externally to the direction of travel (37 degrees and 21 degrees , respectively) so that pure flexion and extension at the knee will act to adduct and adbuct the tibiotarsus relative to the plane of movement, and pure flexion and extension at the ankle will act to abduct and adduct the tarsometatarsus relative to the plane of movement. This feature of the limb anatomy appears to provide the major lateral (non-sagittal) displacement of the lower limb necessary for steering the swinging limb clear of the stance limb and replaces what would otherwise require greater adduction/abduction and/or internal/external rotation, allowing for less complex joints, musculoskeletal geometry and neuromuscular control. Significant rotation about the joints' non-flexion/extension axes nevertheless occurs over the running stride. In particular, hip abduction and knee internal/external and varus/valgus motion may further facilitate limb clearance during the swing phase, and substantial non-flexion/extension movement at the knee is also observed during stance. Measurement of 3-D segment and joint motion in birds will be aided by the use of functionally determined axes of rotation rather than assumed axes, proving important when interpreting the biomechanics and motor control of avian bipedalism.

  11. Anatomy of the proximal tibiofibular joint and interosseous membrane, and their contributions to joint kinematics in below-knee amputations.

    PubMed

    Burkhart, Timothy A; Asa, Benjamin; Payne, Michael W C; Johnson, Marjorie; Dunning, Cynthia E; Wilson, Timothy D

    2015-02-01

    A result of below-knee amputations (BKAs) is abnormal motion that occurs about the proximal tibiofibular joint (PTFJ). While it is known that joint morphology may play a role in joint kinematics, this is not well understood with respect to the PTFJ. Therefore, the purposes of this study were: (i) to characterize the anatomy of the PTFJ and statistically analyze the relationships within the joint; and (ii) to determine the relationships between the PTFJ characteristics and the degree of movement of the fibula in BKAs. The PTFJ was characterized in 40 embalmed specimens disarticulated at the knee, and amputated through the mid-tibia and fibula. Four metrics were measured: inclination angle (angle at which the fibula articulates with the tibia); tibial and fibular articular surface areas; articular surface concavity and shape. The specimens were mechanically tested by applying a load through the biceps femoris tendon, and the degree of motion about the tibiofibular joint was measured. Regression analyses were performed to determine the relationships between the different PTFJ characteristics and the magnitude of fibular abduction. Finally, Pearson correlation analyses were performed on inclination angle and surface area vs. fibular kinematics. The inclination angle measured on the fibula was significantly greater than that measured on the tibia. This difference may be attributed to differences in concavity of the tibial and fibular surfaces. Surface area measured on the tibia and fibula was not statistically different. The inclination angle was not statistically correlated to surface area. However, when correlating fibular kinematics in BKAs, inclination angle was positively correlated to the degree of fibular abduction, whereas surface area was negatively correlated. The characteristics of the PTFJ dictate the amount of fibular movement, specifically, fibular abduction in BKAs. Predicting BKA complications based on PTFJ characteristics can lead to recommendations in treatment. © 2014 Anatomical Society.

  12. An optimal resolved rate law for kindematically redundant manipulators

    NASA Technical Reports Server (NTRS)

    Bourgeois, B. J.

    1987-01-01

    The resolved rate law for a manipulator provides the instantaneous joint rates required to satisfy a given instantaneous hand motion. When the joint space has more degrees of freedom than the task space, the manipulator is kinematically redundant and the kinematic rate equations are underdetermined. These equations can be locally optimized, but the resulting pseudo-inverse solution was found to cause large joint rates in some case. A weighting matrix in the locally optimized (pseudo-inverse) solution is dynamically adjusted to control the joint motion as desired. Joint reach limit avoidance is demonstrated in a kinematically redundant planar arm model. The treatment is applicable to redundant manipulators with any number of revolute joints and to nonplanar manipulators.

  13. The influence of abutment screw tightening on screw joint configuration.

    PubMed

    Lang, Lisa A; Wang, Rui-Feng; May, Kenneth B

    2002-01-01

    Limiting abutment-to-implant hexagonal discrepancies and rotational movement of the abutment around the implant to less than 5 degrees would result in a more stable screw joint. However, the exact relationship after abutment screw tightening is unknown, as is the effect of a counter-torque device in limiting abutment movement during screw tightening. This study examined the orientation of the abutment hexagon to the implant hexagon after tightening of the abutment screw for several abutment systems with and without the use of a counter-torque device. Thirty conical self-tapping implants (3.75 x 10.0 mm) and 10 wide-platform Brånemark System implants (5.0 x 10.0 mm), along with 10 abutment specimens from the CeraOne, Estheticone, Procera, and AuraAdapt systems, were selected for this investigation. The implants were placed in a holding device prior to tightening of the abutments. When the tightening torque recommended for each abutment system was reached with the use of a torque controller, each implant abutment specimen was removed from the holding device and embedded in a hard resin medium. The specimens were sectioned in a horizontal direction at the level of the hexagons and cleansed of debris prior to examination. The hexagon orientations were assessed as the degree and direction of rotation of the abutment hexagon around the implant hexagon. The range of the maximum degrees of rotation for all 4 abutment groups tightened with or without the counter-torque device was slightly more than 3.53 degrees. The absolute degrees of rotation for all 4 abutment groups were less than 1.50 degrees with or without the use of the counter-torque device. The hexagon-to-hexagon orientation measured as rotational fit on all abutment systems was below the 5 degrees suggested as optimal for screw joint stability. The absolute degrees of rotation for all 4 abutment groups were less than 1.50 degrees regardless of whether the counter-torque device was used.

  14. The responses of muscle spindles to small, slow movements in passive muscle and during fusimotor activity.

    PubMed

    Wise, A K; Gregory, J E; Proske, U

    1999-03-06

    We have previously shown that movement detection thresholds at the human elbow joint were less than a degree of joint rotation in the passive limb but were higher if they were measured while subjects co-contracted elbow muscles [A.K. Wise, J.E. Gregory, U. Proske, J. Physiol., 508 (1998) 325-330]. Here we report observations on the responses of muscle spindles of the soleus muscle of the anaesthetised cat to determine their ability to signal small length changes in the passive muscle and during a contraction, under conditions resembling those of the human experiments. After appropriate conditioning of the muscle to control for history effects, primary endings of muscle spindles showed thresholds to ramp stretch at 20 micrometers s-1 of between less than 5 micrometers and 15 micrometers, which translates to 0.05 degrees -0.15 degrees of human elbow joint rotation. Thresholds were much higher following conditioning to introduce slack in the muscle. Since during a voluntary contraction there is likely to be alpha:gamma co-activation, responses of spindles were also recorded during slow stretches (100 micrometers at 20 micrometers s-1) during static fusimotor stimulation, dynamic fusimotor stimulation, combined fusimotor stimulation and fusimotor plus skeletomotor stimulation. Invariably, responses to passive stretch were larger than during motor stimulation. It is concluded that spindles are sensitive enough to signal fractions of a degree of elbow joint rotation and that the rise in threshold observed during a voluntary contraction may be accounted for by the actions of fusimotor and skeletomotor axons on spindle stretch responses. Copyright 1999 Elsevier Science B.V.

  15. Three new bachelors of photonics in Ontario

    NASA Astrophysics Data System (ADS)

    Nantel, Marc; Beda, Johann; Grevatt, Treena; Chebbi, Brahim; Jessop, Paul; Song, Shaowen

    2004-10-01

    After the introduction in 2001 of community college programs at the Photonics Technician/Technologist levels, the need to cover the photonics educational space at the undergraduate level was addressed. In the last year, three very different new undergraduate degrees in photonics have started to develop in Ontario. These programs are presented in this paper. The Honours B.Sc. in Photonics at Wilfrid Laurier University (Waterloo) will develop a strong understanding of the theory and application of photonics, with practical hands-on exposure to optics, fibre optics, and lasers. This program benefits from the particularity that the department offering it combines both Physics and Computer Science. At McMaster University, the Engineering Physics program will provide students with a broad background in basic Engineering, Mathematics, Electronics, and Semiconductors, as well as an opportunity to pursue Photonics in greater depth and to have that fact recognized in the program designation. The Niagara and Algonquin College Bachelor of Applied Technology in Photonics program is co-op and joint between the two institutions. Emphasis is placed on the applied aspects of the field, with the more hands-on experimental learning taking precedence in the first years and the more advanced theoretical subjects following in the latter years.

  16. Smaller external notebook mice have different effects on posture and muscle activity.

    PubMed

    Oude Hengel, Karen M; Houwink, Annemieke; Odell, Dan; van Dieën, Jaap H; Dennerlein, Jack T

    2008-07-01

    Extensive computer mouse use is an identified risk factor for computer work-related musculoskeletal disorders; however, notebook computer mouse designs of varying sizes have not been formally evaluated but may affect biomechanical risk factors. Thirty adults performed a set of mouse tasks with five notebook mice, ranging in length from 75 to 105 mm and in width from 35 to 65 mm, and a reference desktop mouse. An electro-magnetic motion analysis system measured index finger (metacarpophalangeal joint), wrist and forearm postures, and surface electromyography measured muscle activity of three extensor muscles in the forearm and the first dorsal interosseus. The smallest notebook mice were found to promote less neutral postures (up to 3.2 degrees higher metacarpophalangeal joint adduction; 6.5 degrees higher metacarpophalangeal joint flexion, 2.3 degrees higher wrist extension) and higher muscle activity (up to 4.1% of maximum voluntary contraction higher wrist extensor muscle activity). Participants with smaller hands had overall more non-neutral postures than participants with larger hands (up to 5.6 degrees higher wrist extension and 5.9 degrees higher pronation); while participants with larger hands were more influenced by the smallest notebook mice (up to 3.6 degrees higher wrist extension and 5.5% of maximum voluntary contraction higher wrist extensor values). Self-reported ratings showed that while participants preferred smaller mice for portability; larger mice scored higher on comfort and usability. The smallest notebook mice increased the intensity of biomechanical exposures. Longer term mouse use could enhance these differences, having a potential impact on the prevention of work-related musculoskeletal disorders.

  17. Atlas 5013 tank corrosion test

    NASA Technical Reports Server (NTRS)

    Sutherland, W. M.; Girton, L. D.; Treadway, D. G.

    1978-01-01

    The type and cause of corrosion in spot welded joints were determined by X-ray and chemical analysis. Fatigue and static tests showed the degree of degradation of mechanical properties. The corrosion inhibiting effectiveness of WD-40 compound and required renewal period by exposing typical joint specimens were examined.

  18. L'espace articulaire de la Robotique Industrielle est un espace vectorielIndustrial Robotics joint space is a vector space

    NASA Astrophysics Data System (ADS)

    Tondu, Bertrand

    2003-05-01

    The mathematical modelling of industrial robots is based on the vectorial nature of the n-dimensional joint space of the robot, defined as a kinematic chain with n degrees of freedom. However, in our opinion, the vectorial nature of the joint space has been insufficiently discussed in the literature. We establish the vectorial nature of the joint space of an industrial robot from the fundamental studies of B. Roth on screws. To cite this article: B. Tondu, C. R. Mecanique 331 (2003).

  19. 75 FR 82132 - ITS Joint Program Office; Human Factors for IntelliDrive SM

    Federal Register 2010, 2011, 2012, 2013, 2014

    2010-12-29

    ... DEPARTMENT OF TRANSPORTATION ITS Joint Program Office; Human Factors for IntelliDrive \\SM\\ (HFID... meeting to discuss the Human Factors for IntelliDrive (HFID) program on January 6, 2011 from 10 a.m. to 5... environment. One of the supporting research programs is Human Factors for IntelliDrive (HFID), a program aimed...

  20. [CT measurement and clinical application of double-row suture anchor reconstruction for the treatment of Tossy type III acromioclavicular joint dislocation].

    PubMed

    Zhang, Chuan-Kai; Liu, Chen; Han, Bing; Feng, Hui; Chen, Qi-Zhong; Sunx, Sun Yi-Yan

    2017-04-25

    To study feasibility and reliability of reconstruction of the acromioclavicular ligament with double-row suture anchor for the treatment of acromioclavicular joint dislocation through coracoid coronal CT measurement, and to provide a new operation method for treating acromioclavicular joint dislocation. Total 60 healthy people received CT examination of shoulder joint, including 30 males and 30 females, ranging in age from 18 to 50 years old. The coronal width, thickness and 20 degree camber angle in the medial part of the toot of coronal were measured using CT scan. The results were applied to clinical treatment for 12 patients with acromioclavicular joint dislocation of Tossy III type. The width in the medial part of the root of the coracoid was(17.65±1.82) mm(left side) and (17.67±1.80) mm(right side) in males; (16.55±1.78) mm(left side) and (16.52±1.74) mm (right side) in females. The vertical thickness of the roots of the coracoid: (13.11±2.11) mm(left side) and (13.16±2.09) mm(right side) in males;(12.79±2.21) mm(left side) and (12.76±2.19) mm (right side) in females. The thickness of 20 degrees camber angle of the coracoid roots: (16.32±1.74) mm (left side) and (16.30±1.69) mm(right side) in males; (15.68±1.44) mm(left side) and (15.67±1.43) mm(right side) in females. Total 12 patients were treated with anchor nail with extraversion 20 degrees. The postoperative X-ray films showed bone anchors were located in the coracoid process, no bone splitting. Double-row suture anchor of 5 mm diameter nails can be placed into coracoid with extraversion 20 degrees, which is safety.

  1. Joint Program Management Handbook

    DTIC Science & Technology

    1994-12-01

    service networks , and unique requirements, such as in the special opeations area, require a special effort by joint program managers Figure 3-3...the Engieermg and Manufacuring Devopment Phase. Nfilestoae HI- Develommen Annros Devopment approval marks a significant step for any program, but it is

  2. Tribute to Dr Jacques Rogge: muscle activity and fatigue during hiking in Olympic dinghy sailing.

    PubMed

    Bourgois, Jan G; Dumortier, Jasmien; Callewaert, Margot; Celie, Bert; Capelli, Carlo; Sjøgaard, Gisela; De Clercq, Dirk; Boone, Jan

    2017-06-01

    'A tribute to Dr J. Rogge' aims to systematically review muscle activity and muscle fatigue during sustained submaximal quasi-isometric knee extension exercise (hiking) related to Olympic dinghy sailing as a tribute to Dr Rogge's merits in the world of sports. Dr Jacques Rogge is not only the former President of the International Olympic Committee, he was also an orthopaedic surgeon and a keen sailor, competing at three Olympic Games. In 1972, in fulfilment of the requirements for the degree of Master in Sports Medicine, he was the first who studied a sailors' muscle activity by means of invasive needle electromyography (EMG) during a specific sailing technique (hiking) on a self-constructed sailing ergometer. Hiking is a bilateral and multi-joint submaximal quasi-isometric movement which dinghy sailors use to optimize boat speed and to prevent the boat from capsizing. Large stresses are generated in the anterior muscles that cross the knee and hip joint, mainly employing the quadriceps at an intensity of 30-40% maximal voluntary contraction (MVC), sometimes exceeding 100% MVC. Better sailing level is partially determined by a lower rate of neuromuscular fatigue during hiking and for ≈60% predicted by a higher maximal isometric quadriceps strength. Although useful in exercise testing, prediction of hiking endurance capacity based on the changes in surface EMG in thigh and trunk muscles during a hiking maintenance task is not reliable. This could probably be explained by the varying exercise intensity and joint angles, and the great number of muscles and joints involved in hiking. Highlights Dr Jacques Rogge, former president of the International Olympic Committee and Olympic Finn sailor, was the first to study muscle activity during sailing using invasive needle EMG to obtain his Master degree in Sports Medicine at the Ghent University. Hiking is a critical bilateral and multi-joint movement during dinghy racing, accounting for >60% of the total upwind leg time. Hiking generates large stresses in the anterior muscles that cross the knee and hip joint. Hiking is considered as a quasi-isometric bilateral knee extension exercise. Muscle activity measurements during sailing, recorded by means of EMG, show a mean contraction intensity of 30-40% maximal voluntary contraction with peaks exceeding 100%. Hiking performance is strongly related to the development of neuromuscular fatigue in the quadriceps muscle. Since maximal strength is an important determinant of neuromuscular fatigue during hiking, combined strength and endurance training should be incorporated in the training program of dinghy sailors.

  3. Metal to ceramic sealed joint

    DOEpatents

    Lasecki, J.V.; Novak, R.F.; McBride, J.R.

    1991-08-27

    A metal to ceramic sealed joint which can withstand wide variations in temperature and maintain a good seal is provided for use in a device adapted to withstand thermal cycling from about 20 to about 1000 degrees C. The sealed joint includes a metal member, a ceramic member having an end portion, and an active metal braze forming a joint to seal the metal member to the ceramic member. The joint is positioned remote from the end portion of the ceramic member to avoid stresses at the ends or edges of the ceramic member. The sealed joint is particularly suited for use to form sealed metal to ceramic joints in a thermoelectric generator such as a sodium heat engine where a solid ceramic electrolyte is joined to metal parts in the system. 11 figures.

  4. Metal to ceramic sealed joint

    DOEpatents

    Lasecki, John V.; Novak, Robert F.; McBride, James R.

    1991-01-01

    A metal to ceramic sealed joint which can withstand wide variations in temperature and maintain a good seal is provided for use in a device adapted to withstand thermal cycling from about 20 to about 1000 degrees C. The sealed joint includes a metal member, a ceramic member having an end portion, and an active metal braze forming a joint to seal the metal member to the ceramic member. The joint is positioned remote from the end portion of the ceramic member to avoid stresses at the ends or edges of the ceramic member. The sealed joint is particularly suited for use to form sealed metal to ceramic joints in a thermoelectric generator such as a sodium heat engine where a solid ceramic electrolyte is joined to metal parts in the system.

  5. Kinematics of the six-degree-of-freedom force-reflecting Kraft Master

    NASA Technical Reports Server (NTRS)

    Williams, Robert L., II

    1991-01-01

    Presented here are kinematic equations for a six degree of freedom force-reflecting hand controller. The forward kinematics solution is developed and shown in simplified form. The Jacobian matrix, which uses terms from the forward kinematics solution, is derived. Both of these kinematic solutions require joint angle inputs. A calibration method is presented to determine the hand controller joint angles given the respective potentiometer readings. The kinematic relationship describing the mechanical coupling between the hand and controller shoulder and elbow joints is given. These kinematic equations may be used in an algorithm to control the hand controller as a telerobotic system component. The purpose of the hand controller is two-fold: operator commands to the telerobotic system are entered using the hand controller, and contact forces and moments from the task are reflected to the operator via the hand controller.

  6. Automatically Locking/Unlocking Orthotic Knee Joint

    NASA Technical Reports Server (NTRS)

    Weddendorf, Bruce

    1994-01-01

    Proposed orthotic knee joint locks and unlocks automatically, at any position within range of bend angles, without manual intervention by wearer. Includes tang and clevis, locks whenever wearer transfers weight to knee and unlocks when weight removed. Locking occurs at any angle between 45 degrees knee bend and full extension.

  7. The Joint NASA/Goddard-University of Maryland Research Program in Charged Particle and High Energy Photon Detector Technology

    NASA Technical Reports Server (NTRS)

    1988-01-01

    Having recognized at an early stage the critical importance of maintaining detector capabilities which utilize state of the art techniques, a joint program was formulated. This program has involved coordination of a broad range of efforts and activities including joint experiments, collaboration in theoretical studies, instrument design, calibrations, and data analysis. Summaries of the progress made to date are presented. A representative bibliography is also included.

  8. Magic-Angle-Spinning NMR Magnet Development: Field Analysis and Prototypes

    PubMed Central

    Voccio, John; Hahn, Seungyong; Park, Dong Keun; Ling, Jiayin; Kim, Youngjae; Bascuñán, Juan; Iwasa, Yukikazu

    2013-01-01

    We are currently working on a program to complete a 1.5 T/75 mm RT bore magic-angle-spinning nuclear magnetic resonance magnet. The magic-angle-spinning magnet comprises a z-axis 0.866-T solenoid and an x-axis 1.225-T dipole, each to be wound with NbTi wire and operated at 4.2 K in persistent mode. A combination of the fields creates a 1.5-T field pointed at 54.74 degrees (magic angle) from the rotation (z) axis. In the first year of this 3-year program, we have completed magnetic analysis and design of both coils. Also, using a winding machine of our own design and fabrication, we have wound several prototype dipole coils with NbTi wire. As part of this development, we have repeatedly made successful persistent NbTi-NbTi joints with this multifilamentary NbTi wire. PMID:24058275

  9. Development and Implementation of Joint Programs in Laser Ranging and Other Space Geodetic Techniques

    NASA Technical Reports Server (NTRS)

    Pearlman, Michael R.; Carter, David (Technical Monitor)

    2004-01-01

    This progress report discusses the status and progress made in joint international programs including: 1) WEGENER; 2) Arabian Peninsula program; 3) Asia-Pacific Space Geodynamics (APSG) program; 4) the Fourteenth International Workshop on Laser Ranging; 5) the International Laser Ranging Service; and 6) current support for the NASA network.

  10. In vivo six-degree-of-freedom knee-joint kinematics in overground and treadmill walking following total knee arthroplasty.

    PubMed

    Guan, Shanyuanye; Gray, Hans A; Schache, Anthony G; Feller, Julian; de Steiger, Richard; Pandy, Marcus G

    2017-08-01

    No data are available to describe six-degree-of-freedom (6-DOF) knee-joint kinematics for one complete cycle of overground walking following total knee arthroplasty (TKA). The aims of this study were firstly, to measure 6-DOF knee-joint kinematics and condylar motion for overground walking following TKA; and secondly, to determine whether such data differed between overground and treadmill gait when participants walked at the same speed during both tasks. A unique mobile biplane X-ray imaging system enabled accurate measurement of 6-DOF TKA knee kinematics during overground walking by simultaneously tracking and imaging the joint. The largest rotations occurred for flexion-extension and internal-external rotation whereas the largest translations were associated with joint distraction and anterior-posterior drawer. Strong associations were found between flexion-extension and adduction-abduction (R 2  = 0.92), joint distraction (R 2  = 1.00), and anterior-posterior translation (R 2  = 0.77), providing evidence of kinematic coupling in the TKA knee. Although the measured kinematic profiles for overground walking were grossly similar to those for treadmill walking, several statistically significant differences were observed between the two conditions with respect to temporo-spatial parameters, 6-DOF knee-joint kinematics, and condylar contact locations and sliding. Thus, caution is advised when making recommendations regarding knee implant performance based on treadmill-measured knee-joint kinematic data. © 2016 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 35:1634-1643, 2017. © 2016 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.

  11. Common Bolted Joint Analysis Tool

    NASA Technical Reports Server (NTRS)

    Imtiaz, Kauser

    2011-01-01

    Common Bolted Joint Analysis Tool (comBAT) is an Excel/VB-based bolted joint analysis/optimization program that lays out a systematic foundation for an inexperienced or seasoned analyst to determine fastener size, material, and assembly torque for a given design. Analysts are able to perform numerous what-if scenarios within minutes to arrive at an optimal solution. The program evaluates input design parameters, performs joint assembly checks, and steps through numerous calculations to arrive at several key margins of safety for each member in a joint. It also checks for joint gapping, provides fatigue calculations, and generates joint diagrams for a visual reference. Optimum fastener size and material, as well as correct torque, can then be provided. Analysis methodology, equations, and guidelines are provided throughout the solution sequence so that this program does not become a "black box:" for the analyst. There are built-in databases that reduce the legwork required by the analyst. Each step is clearly identified and results are provided in number format, as well as color-coded spelled-out words to draw user attention. The three key features of the software are robust technical content, innovative and user friendly I/O, and a large database. The program addresses every aspect of bolted joint analysis and proves to be an instructional tool at the same time. It saves analysis time, has intelligent messaging features, and catches operator errors in real time.

  12. The effects of transcutaneous electrical nerve stimulation on joint position sense in patients with knee joint osteoarthritis.

    PubMed

    Shirazi, Zahra Rojhani; Shafaee, Razieh; Abbasi, Leila

    2014-10-01

    To study the effects of transcutaneous electrical nerve stimulation (TENS) on joint position sense (JPS) in knee osteoarthritis (OA) subjects. Thirty subjects with knee OA (40-60 years old) using non-random sampling participated in this study. In order to evaluate the absolute error of repositioning of the knee joint, Qualysis Track Manager system was used and sensory electrical stimulation was applied through the TENS device. The mean errors in repositioning of the joint, in two position of the knee joint with 20 and 60 degree angle, after applying the TENS was significantly decreased (p < 0.05). Application of TENS in subjects with knee OA could improve JPS in these subjects.

  13. An optimal resolved rate law for kinematically redundant manipulators

    NASA Technical Reports Server (NTRS)

    Bourgeois, B. J.

    1987-01-01

    The resolved rate law for a manipulator provides the instantaneous joint rates required to satisfy a given instantaneous hand motion. When the joint space has more degrees of freedom than the task space, the manipulator is kinematically redundant and the kinematic rate equations are underdetermined. These equations can be locally optimized, but the resulting pseudo-inverse solution has been found to cause large joint rates in some cases. A weighting matrix in the locally optimized (pseudo-inverse) solution is dynamically adjusted to control the joint motion as desired. Joint reach limit avoidance is demonstrated in a kinematically redundant planar arm model. The treatment is applicable to redundant manipulators with any number of revolute joints and to non-planar manipulators.

  14. Quiet Short-Haul Research Aircraft Joint Navy/NASA Sea Trials

    NASA Technical Reports Server (NTRS)

    Queen, S.; Cochrane, J.

    1982-01-01

    The Quiet Short-Haul Research Aircraft (QSRA) is a flight facility which Ames Research Center is using to conduct a broad program of terminal area and low-speed, propulsive-life flight research. A joint Navy/NASA flight research program used the QSRA to investigate the application of advanced propulsive-lift technology to the naval aircraft-carrier environment. Flight performance of the QSRA is presented together with the results or the joint Navy/NASA flight program. During the joint program, the QSRA operated aboard the USS Kitty Hawk for 4 days, during which numerous unarrested landings and free deck takeoffs were accomplished. These operations demonstrated that a large aircraft incorporating upper-surface-blowing, propulsive-life technology can be operated in the aircraft-carrier environment without any unusual problems.

  15. Discrete and continuous joint coupling relationships in uninjured recreational runners.

    PubMed

    Dierks, Tracy A; Davis, Irene

    2007-06-01

    Abnormal joint coupling is thought to be related to overuse injuries in runners. However, researchers do not yet know what constitutes normal joint coupling during running, which makes abnormal coupling difficult to define. Lower extremity kinematics were collected from 40 recreational runners during stance. Joint coupling methods were applied and, for each method, means and both within- and between-subject variability were calculated. The 95% confidence interval was used to compare differences across coupling relationships and periods of stance. Timing between rearfoot eversion, tibial internal rotation, and knee flexion were relatively synchronous while relationships involving knee internal rotation were more asynchronous. The excursion ratios showed that every 2 degrees of rearfoot eversion was coupled with 1 degrees of both tibial internal rotation and knee internal rotation. Vector coding results showed that just beyond maximum loading, all joint coupling relationships resulted in relatively equal amounts of motion, while the within-subject variability was similar throughout stance. The continuous relative phase results showed that the most out-of-phase coupling occurred in the periods around heel-strike and toe-off while the most in-phase coupling occurred in the period just beyond maximum loading of the leg. The continuous relative phase within-subject variability was greatest at the periods around heel-strike and toe-off and smallest just beyond maximum loading. With a better understanding of joint coupling in uninjured runners, these data will help to serve as a reference for future studies investigating the relationship between running injuries and abnormal joint coupling.

  16. Efficacy of Caregiver-Mediated Joint Engagement Intervention for Young Children with Autism Spectrum Disorders

    ERIC Educational Resources Information Center

    Chiang, Chung-Hsin; Chu, Ching-Lin; Lee, Tsung-Chin

    2016-01-01

    Joint attention intervention for children with autism spectrum disorders was focused on improving joint engagement and joint attention skills. The purpose of this study was to develop a caregiver-mediated joint engagement intervention program combined with body movement play to investigate the effects of joint engagement/joint attention skills in…

  17. [Treatment of pediatric distal femur fractures by external fixator combined with limited internal fixation].

    PubMed

    Wei, Sheng-wang; Shi, Zhan-ying; Hu, Ju-zheng; Wu, Hao

    2016-03-01

    To discuss the clinical effects of external fixator combined with limited internal fixation in the treatment of pediatric distal femur fractures. From January 2008 to June 2014, 17 children of distal femur fractures were treated by external fixator combined with limited internal fixation. There were 12 males and 5 females, aged from 6 to 13 years old with an average of 10.2 years, ranged in the course of disease from 1 h to 2 d. Preoperative diagnoses were confirmed by X-ray films in all children. There were 11 patients with supracondylar fracture , and 6 patients with intercondylar comminuted fracture. According to AO/ASIF classification, 9 fractures were type A1, 5 cases were type A2,and 3 cases were type C1. The intraoperative and postoperative complications, postoperative radiological examination, lower limbs length and motion of knee joints were observed. Knee joint function was assessed by KSS score. All the patients were followed up from 6 to 38 months with an average of 24.4 months. No nerve or blood vessel injury was found. One case complicated with the external fixation loosening, 2 cases with the infection of pin hole and 3 cases with the leg length discrepancy. Knee joint mobility and length measurement (compared with the contralateral), the average limited inflexion was 10 degrees (0 degrees to 20 degrees), the average limited straight was 4 degrees (0 degrees to 10), the average varus or valgus angle was 3 degrees (0 degrees to 5 degrees). KSS of the injured side was (96.4 +/- 5.0) points at final follow-up, 16 cases got excellent results and 1 good. All fractures obtained healing and no epiphyseal closed early was found. External fixator combined with limited internal fixation has advantages of simple operation, reliable fixation, early functional exercise in treating pediatric distal femurs fractures.

  18. Aerodynamic performance and pressure distributions for a NASA SC(2)-0714 airfoil tested in the Langley 0.3-meter transonic cryogenic tunnel

    NASA Technical Reports Server (NTRS)

    Jenkins, Renaldo V.; Hill, Acquilla S.; Ray, Edward J.

    1988-01-01

    This report presents in graphic and tabular forms the aerodynamic coefficient and surface pressure distribution data for a NASA SC(2)-0714 airfoil tested in the Langley 0.3-Meter Transonic Cryogenic Tunnel. The test was another in a series of tests involved in the joint NASA/U.S. Industry Advanced Technology Airfoil Tests program. This 14% thick supercritical airfoil was tested at Mach numbers from 0.6 to 0.76 and angles of attack from -2.0 to 6.0 degrees. The test Reynolds numbers were 4 million, 6 million, 10 million, 15 million, 30 million, 40 million, and 45 million.

  19. [Investigation of body weight ratios on joint structures at different knee flexion angles in patients with unilateral knee arthroplasty].

    PubMed

    Bakırhan, Serkan; Unver, Bayram; Karatosun, Vasfi

    2013-01-01

    The study aims to determine body weight ratios between extremities in patients with unilateral total knee arthroplasty (TKA) at 12 months postoperatively at the static-standing position at 30, 60 and 90 degrees of knee flexion. The study included 52 female patients (mean age 65.6±10.6 years; range 40 to 83 years) who underwent unilateral primary TKA. The force-platform was used to calculate the body-weight ratios of the patients. Body weight ratios on the operated and non-operated limbs of the unilateral TKA patients were examined at standing-static position at 30, 60 and 90 degrees of knee flexion on the force-platform according to their age and body mass index (BMI). The pain levels of the patients were evaluated using the visual analog scale. It was found that unilateral TKA patients placed their body weight on the non-operated limb more at the standing-static position, and 30, 60 and 90 degrees of knee flexion at 12 months postoperatively (p<0.05). It was also found that as the knee flexion degree increased with age, so did TKA patients place their body weight on the nonoperated limb more (p<0.05), and that BMI had no effect on the load distribution difference over the two extremities (p>0.05). During the postoperative period, load asymmetry between the two extremities in patients with unilateral TKA remains the same due to advancing age. This accelerates the osteoarthritis process on the non-operated knee. It is concluded that the age factor should be taken into account while planning physiotherapy and rehabilitation programs for unilateral TKA patients and knee exercise programs aiming to place load over the operated limb should be arranged.

  20. SEMTAP (Serpentine End Match TApe program): The Easy Way to Program Your Numerically Controlled Router for the Production of SEM Joints

    Treesearch

    Ronald E. Coleman

    1977-01-01

    SEMTAP (Serpentine End Match TApe Program) is an easy and inexpensive method of programing a numerically controlled router for the manufacture of SEM (Serpentine End Matching) joints. The SEMTAP computer program allows the user to issue commands that will accurately direct a numerically controlled router along any SEM path. The user need not be a computer programer to...

  1. Critical joints in large composite aircraft structure

    NASA Technical Reports Server (NTRS)

    Nelson, W. D.; Bunin, B. L.; Hart-Smith, L. J.

    1983-01-01

    A program was conducted at Douglas Aircraft Company to develop the technology for critical structural joints of composite wing structure that meets design requirements for a 1990 commercial transport aircraft. The prime objective of the program was to demonstrate the ability to reliably predict the strength of large bolted composite joints. Ancillary testing of 180 specimens generated data on strength and load-deflection characteristics which provided input to the joint analysis. Load-sharing between fasteners in multirow bolted joints was computed by the nonlinear analysis program A4EJ. This program was used to predict strengths of 20 additional large subcomponents representing strips from a wing root chordwise splice. In most cases, the predictions were accurate to within a few percent of the test results. In some cases, the observed mode of failure was different than anticipated. The highlight of the subcomponent testing was the consistent ability to achieve gross-section failure strains close to 0.005. That represents a considerable improvement over the state of the art.

  2. Replacement of seam welded hot reheat pipe using narrow groove GTA machine welding

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Richardson, R.R.; Yanes, J.; Bryant, R.

    1995-12-31

    Southern California Edison, recognizing a potential safety concern, scrutinized its existing seam welded hot reheat pipe manufactured by the same supplier as that which failed. Alternatives were narrowed to two in dealing with the installed seam welded pipe. The overriding consideration, however, was one of safety. With this in mind, the utility company evaluated replacement of the seam welded hot reheat pipe with seamless pipe or increasing the frequency of its inspection program. Although increased inspection was much costly, pipe replacement was chosen due to potential safety concerns with seam welded pipe even with more frequent inspection. The utility companymore » then proceeded to determine the most effective method to complete this work. Analysis showed machine-made (automatic) gas tungsten arc welds (GTAW) as the method of choice due to cleanliness and superior mechanical properties. In conjunction with this method, the narrow groove (3{degree} bevel) weld joint as opposed to the traditional groove (37 1/2{degree} bevel) was shown to provide significant technical advantages.« less

  3. Joint University Program for Air Transportation Research, 1984

    NASA Technical Reports Server (NTRS)

    Morrell, Frederick R. (Compiler)

    1987-01-01

    The research conducted during 1984 under the NASA/FAA sponsored Joint University Program for Air Transportation Research is summarized. The Joint University Program is a coordinated set of three grants sponsored by NASA Langley Research Center and the Federal Aviation Administration, one each with the Massachusetts Institute of Technology, Ohio University, and Princeton University. Completed works, status reports, and bibliographies are presented for research topics, which include navigation, guidance, control and display concepts. An overview of the year's activities for each of the schools is also presented.

  4. Joint University Program for Air Transportation Research, 1986

    NASA Technical Reports Server (NTRS)

    Morrell, Frederick R. (Compiler)

    1988-01-01

    The research conducted under the NASA/FAA sponsored Joint University Program for Air Transportation Research is summarized. The Joint University Program is a coordinated set of three grants sponsored by NASA and the FAA, one each with the Mass. Inst. of Tech., Ohio Univ., and Princeton Univ. Completed works, status reports, and bibliographies are presented for research topics, which include computer science, guidance and control theory and practice, aircraft performance, flight dynamics, and applied experimental psychology. An overview of activities is presented.

  5. Regional Security in the Gulf of Guinea; A Vital Interest of the United States

    DTIC Science & Technology

    2010-03-18

    Davis Highway, Suite 1204, Arlington VA 22202-4302. Respondents should be aware that notwithstanding any other provision of law , no person shall be...A paper submitted to the Faculty of the Joint Advanced Warfighting School in partial satisfaction of the requirements of a Master of Science...Degree in Joint Campaign Planning and Strategy. The contents of this paper reflect my personal views and are not necessarily endorsed by the Joint

  6. Managing Non-Standard Force Demands: Risk Implications of the Global Force Management System

    DTIC Science & Technology

    2012-04-26

    GLOBAL FORCE MANAGEMENT SYSTEM by James C. Wright GS-14, Department of Defense A paper submitted to the Faculty of the Joint Advanced Warfighting...School in partial satisfaction of the requirements of a Master of Science Degree in Joint Campaign Planning and Strategy. The contents of this paper ...reflect my own personal views and are not necessarily endorsed by the Joint Forces Staff College or the Department of Defense. This paper is entirely

  7. Arthrometric Evaluation of Stabilizing Effect of Knee Functional Bracing at Different Flexion Angles

    PubMed Central

    Seyed Mohseni, Saeedeh; Moss, Farzam; Karimi, Hossein; Kamali, Mohammad

    2009-01-01

    Previous in-vivo investigations on the stabilizing efficacy of knee bracing for ACL reconstructed patients have been often limited to 20-30 degrees of knee flexion. In this study, the effectiveness of a uniaxial hinged functional brace to improve the knee stability was assessed at 30, 60 and 90 degrees of knee flexion. Arthrometry tests were conducted on 15 healthy subjects before and following wearing the brace and the tibial displacements were measured at up to 150 N anterior forces. Results indicated that functional bracing has a significant stabilizing effect throughout the range of knee flexion examined (p < 0.05). The rate of effectiveness, however, was not consistent across the flexion range, e.g., 50% at 30 degrees and only 4% at 90 degrees. It was suggested that accurate sizing and fitting as well as attention to correct hinge placement relative to the femoral condyles can limit brace migration and improve its effectiveness in mid and deep knee flexion. With using adaptive limb fittings, through flexible pads, and a polycentric joint a more significant improvement of the overall brace performance and efficacy might be obtained. Key points Functional bracing improves the knee joint stability mostly in extension posture. Unlike the non-braced condition, the least knee joint stability appears in mid and deep flexion angles when using a hinged brace. Accurate sizing and fitting and attention to correct hinge placement relative to the femoral condyles can limit brace migration and improve its effectiveness in mid and deep knee flexion. The overall brace performance and efficacy might be improved significantly using adaptive limb fittings through flexible pads and/or polycentric joints. PMID:24149533

  8. Interventions for treating osteoarthritis of the big toe joint.

    PubMed

    Zammit, Gerard V; Menz, Hylton B; Munteanu, Shannon E; Landorf, Karl B; Gilheany, Mark F

    2010-09-08

    Osteoarthritis affecting of the big toe joint of the foot (hallux limitus or rigidus) is a common and painful condition. Although several treatments have been proposed, few have been adequately evaluated. To identify controlled trials evaluating interventions for osteoarthritis of the big toe joint and to determine the optimum intervention(s). Literature searches were conducted across the following electronic databases: CENTRAL; MEDLINE; EMBASE; CINAHL; and PEDro (to 14th January 2010). No language restrictions were applied. Randomised controlled trials, quasi-randomised trials, or controlled clinical trials that assessed treatment outcomes for osteoarthritis of the big toe joint. Participants of any age or gender with osteoarthritis of the big toe joint (defined either radiographically or clinically) were included. Two authors examined the list of titles and abstracts identified by the literature searches. One content area expert and one methodologist independently applied the pre-determined inclusion and exclusion criteria to the full text of identified trials. To minimise error and reduce potential bias, data were extracted independently by two content experts. Only one trial satisfactorily fulfilled the inclusion criteria and was included in this review. This trial evaluated the effectiveness of two physical therapy programs in 20 individuals with osteoarthritis of the big toe joint. Assessment outcomes included pain levels, big toe joint range of motion and plantar flexion strength of the hallux. Mean differences at four weeks follow up were 3.80 points (95% CI 2.74 to 4.86) for self reported pain, 28.30 degrees (95% CI 21.37 to 35.23) for big toe joint range of motion, and 2.80 kg (95% CI 2.13 to 3.47) for muscle strength. Although differences in outcomes between treatment and control groups were reported, the risk of bias was high. The trial failed to employ appropriate randomisation or adequate allocation concealment, used a relatively small sample and incorporated a short follow up (four weeks). No adverse reactions were reported. The reviewed trial presented a high risk of bias, which limited conclusions that could be drawn from the presented data. The inclusion of only one trial indicates the need for more robust randomised controlled trials to determine the efficacy of interventions for this condition.

  9. Biomechanical characteristics of hemi-hamate reconstruction versus volar plate arthroplasty in the treatment of dorsal fracture dislocations of the proximal interphalangeal joint.

    PubMed

    Tyser, Andrew R; Tsai, Michael A; Parks, Brent G; Means, Kenneth R

    2015-02-01

    To compare stability and range of motion after hemi-hamate reconstruction versus volar plate arthroplasty in a biomechanical proximal interphalangeal (PIP) joint fracture-dislocation model. Eighteen digits from 6 cadaver hands were tested. We created defects of 40%, 60%, and 80% in the palmar base of each digit's middle phalanx, simulating an acute PIP joint fracture-dislocation. Each defect scenario was reconstructed with a hemi-hamate arthroplasty followed by a volar plate arthroplasty. A computer-controlled mechanism was used to bring each digit's PIP joint from full extension to full flexion via the digital tendons in each testing state, and in the intact state. During each testing scenario we collected PIP joint cinedata in a true lateral projection using mini-fluoroscopy. A digital radiography program was used to measure the amount of middle phalanx dorsal translation (subluxation) in full PIP joint extension. We recorded the angle at which subluxation, if present, occurred during each testing scenario. Average dorsal displacement of the middle phalanx in relation to the proximal phalanx was 0.01 mm for the hemi-hamate reconstructed joints and -0.03 mm for the volar plate arthroplasty, compared with the intact state. Flexion contractures were noted in each of the specimens reconstructed with volar plate arthroplasty. Degree of contracture was directly correlated with defect size, averaging 20° for 40% defects, 35° for 60% defects, and 60° for 80% defects. We observed no flexion contractures in the hemi-hamate reconstructions. Surgeons can use both hemi-hamate and volar plate arthroplasty to restore PIP joint stability following a fracture dislocation with a large middle phalanx palmar base defect. Use of volar plate arthroplasty led to an increasing flexion contracture as the middle phalanx palmar base defect increased. Clinicians can use the information from this study to help with surgical decision-making and patient education. Copyright © 2015 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

  10. Influence of Joint Flexibility on Vibration Analysis of Free-Free Beams

    NASA Astrophysics Data System (ADS)

    Gunda, Jagadish Babu; Krishna, Y.

    2014-12-01

    In present work, joint flexibility (or looseness) of the free-free beam is investigated by using a two noded beam finite element formulation with transverse displacement and joint rotations as the degrees of freedom per node at joint location. Flexibility of the joint is primarily represented by means of a rotational spring analogy, where the stiffness of the rotational spring characterizes the looseness of the flexible joint for an applied bending moment. Influence of joint location as well as joint stiffness on modal behavior of first five modes of slender, uniform free-free beams are discussed for various values of non-dimensional rotational spring stiffness parameter. Numerical accuracy of the results obtained from the present finite element formulation are validated by using the commercially available finite element software which shows the confidence gained on the numerical results discussed in the present study.

  11. Design of a knee joint mechanism that adapts to individual physiology.

    PubMed

    Jiun-Yih Kuan; Pasch, Kenneth A; Herr, Hugh M

    2014-01-01

    This paper describes the design of a new knee joint mechanism, called the Adaptive Coupling Joint (ACJ). The new mechanism has an adaptive trajectory of the center of rotations (COR) that automatically matches those of the attached biological joint. The detailed design is presented as well as characterization results of the ACJ. Conventional exoskeleton and assistive devices usually consider limb joints as a one to three degrees of freedom (DOFs) joint synthesized by multiple one-DOF hinge joints in a single plane. However, the biological joints are complex and usually rotate with respect to a changing COR. As a result, the mismatch between limb joint motion and mechanical interface motion can lead to forces that cause undesired ligament and muscle length changes and internal mechanical changes. These undesired changes contribute to discomfort, as well as to the slippage and sluggish interaction between humans and devices. It is shown that the ACJ can transmit planetary torques from either active or passive devices to the limbs without altering the normal biological joint motion.

  12. Modular multimorphic kinematic arm structure and pitch and yaw joint for same

    DOEpatents

    Martin, H. Lee; Williams, Daniel M.; Holt, W. Eugene

    1989-01-01

    A multimorphic kinematic manipulator arm is provided with seven degrees of freedom and modular kinematic redundancy through identical pitch/yaw, shoulder, elbow and wrist joints and a wrist roll device at the wrist joint, which further provides to the manipulator arm an obstacle avoidance capability. The modular pitch/yaw joints are traction drive devices which provide backlash free operation with smooth torque transmission and enhanced rigidity. A dual input drive arrangement is provided for each joint resulting in a reduction of the load required to be assumed by each drive and providing selective pitch and yaw motions by control of the relative rotational directions of the input drive.

  13. Modular multimorphic kinematic arm structure and pitch and yaw joint for same

    DOEpatents

    Martin, H.L.; Williams, D.M.; Holt, W.E.

    1987-04-21

    A multimorphic kinematic manipulator arm is provided with seven degrees of freedom and modular kinematic redundancy through identical pitch/yaw, shoulder, elbow and wrist joints and a wrist roll device at the wrist joint, which further provides to the manipulator arm an obstacle avoidance capability. The modular pitch/yaw joints are traction drive devices which provide backlash free operation with smooth torque transmission and enhanced rigidity. A dual input drive arrangement is provided for each joint resulting in a reduction of the load required to be assumed by each drive means and providing selective pitch and yaw motions by control of the relative rotational directions of the input drive means. 12 figs.

  14. Excluded from the Institutional Habitus: The Joint Student Experience

    ERIC Educational Resources Information Center

    Weissmann, Elke

    2013-01-01

    This article examines the student experience for a particular cohort, namely the joint honours students, at a post-1992 university in the United Kingdom. These students are enrolled in degree courses that combine two subjects at one university. Little attention has so far been given to such students whose experience is decidedly different from…

  15. The joint measurement entanglement can significantly offset the effect of a noisy channel in teleportation

    NASA Astrophysics Data System (ADS)

    Roa, Luis; Ladrón de Guevara, María L.; Soto-Moscoso, Matias; Catalán, Pamela

    2018-05-01

    In our work we consider the following problem in the context of teleportation: an unknown pure state has to be teleported and there are two laboratories which can perform the task. One laboratory uses a pure non-maximally entangled channel but has a capability of performing the joint measurement on bases with a constrained degree of entanglement; the other lab makes use of a mixed X-state channel but can perform a joint measurement on bases with higher entanglement degrees. We compare the average teleportation fidelity achieved in both cases, finding that the fidelity achieved with the X-state can surpass the obtained with a pure channel, even though the X-state is less entangled than the latter. We find the conditions under which this effect occurs. Our results evidence that the entanglement of the joint measurement plays a role as important as the entanglement of the channel in order to optimize the teleportation process. We include an example showing that the average fidelity of teleportation obtained with a Werner state channel can be greater than that obtained with a Bell state channel.

  16. [Coordination patterns assessed by a continuous measure of joints coupling during upper limb repetitive movements].

    PubMed

    Draicchio, F; Silvetti, A; Ranavolo, A; Iavicoli, S

    2008-01-01

    We analyzed the coordination patterns between elbow, shoulder and trunk in a motor task consisting of reaching out, picking up a cylinder, and transporting it back by using the Dynamical Systems Theory and calculating the continuous relative phase (CRP), a continuous measure of the coupling between two interacting joints. We used an optoelectronic motion analysis system consisting of eight infra-red ray cameras to detect the movements of nine skin-mounted markers. We calculated the root square of the adjusted coefficient of determination, the coefficient of multiple correlation (CMC), in order to investigate the repeatability of the joints coordination. The data confirm that the CNS establishes both synergic (i.e. coupling between shoulder and trunk on the frontal plane) and hierarchical (i.e. coupling between elbow-shoulder-trunk on the horizontal plane) relationships among the available degrees of freedom to overcome the complexity due to motor redundancy. The present study describes a method to investigate the organization of the kinematic degrees of freedom during upper limb multi-joint motor tasks that can be useful to assess upper limb repetitive movements.

  17. Design and preliminary evaluation of an exoskeleton for upper limb resistance training

    NASA Astrophysics Data System (ADS)

    Wu, Tzong-Ming; Chen, Dar-Zen

    2012-06-01

    Resistance training is a popular form of exercise recommended by national health organizations, such as the American College of Sports Medicine (ACSM) and the American Heart Association (AHA). This form of training is available for most populations. A compact design of upper limb exoskeleton mechanism for homebased resistance training using a spring-loaded upper limb exoskeleton with a three degree-of-freedom shoulder joint and a one degree-of-freedom elbow joint allows a patient or a healthy individual to move the upper limb with multiple joints in different planes. It can continuously increase the resistance by adjusting the spring length to train additional muscle groups and reduce the number of potential injuries to upper limb joints caused by the mass moment of inertia of the training equipment. The aim of this research is to perform a preliminary evaluation of the designed function by adopting an appropriate motion analysis system and experimental design to verify our prototype of the exoskeleton and determine the optimal configuration of the spring-loaded upper limb exoskeleton.

  18. Dual US-Europe Graduate Degrees in Volcanology

    NASA Astrophysics Data System (ADS)

    Rose, W. I.; van Wyk Devries, B.; Calder, E. S.; Tibaldi, A.

    2010-12-01

    Michigan Tech, Buffalo, Universite Blaise Pascal and University of Milan Bicocca have formed an educational consortium to offer dual MS degrees in volcanology and geotechniques. Students in the program spend half of their MS in Europe and half in the US and have graduate advisory committees that bridge the Atlantic. The new program combines the expertise of four campuses and give students a broader choice of study options than any one campus can offer, while building an international professional experience. The initiative is funded jointly by the US Department of Education and the Education, Audiovisual and Culture Executive Agency (EACEA) of the European Community. Volcanology and geotechniques are global concerns: the volcanological community is fully globalized, while international consortia now deal with major geotechnical problems. Importantly, both fields require clear appreciation of specific local cultural, social and economical conditions. The new generation of researchers and professionals require international vision, but also the ability to understand local conditions. This masters specifically answers this need. This program will give students a language and cultural training in American English, French and Italian, as well as a wide course choice to meet each individual’s professional requirements. Students benefit from both research and professional approaches, acquiring a sound multidisciplinary profile for an excellent start to their careers. The trained INVOGE masters students will: meet a clear need for professionals/researchers with broad volcanology/geotechniques skills, and provide a workforce with international vision, but capable of addressing local projects. The project is innovative, combining international experience, strong multidisciplinary grounding and a broad subject range: students can choose among many possible advanced coursework and research combinations, and can have a broad choice of graduate advisors, field sites and professional placements. Also, we combine actual mobility with complementary virtual mobility to generate a lasting, coherent joint masters strategy. In the longer term, this exchange of a fraction of our students in both of our strongest research/graduate study areas is expected to add stronger global awareness to all partners, and contribute to improved student preparation, in academic excellence, professional learning and in linguistic/cultural aspects. The goals are to encourage students to consider international universities for PhD work and to increase cross-border professional mobility. This is the first such program in graduate geosciences.

  19. The Effectiveness of a Joint Attention Training Program on Improving Communication Skills of Children with Autism Spectrum Disorder

    ERIC Educational Resources Information Center

    Eissa, Mourad Ali

    2015-01-01

    The purpose of this study was to explore the effect of a joint attention intervention program on improving joint attention and communication skills in children with autism disorder. Participants were ten children between the ages of five and seven who attended a school for children with developmental disabilities (Tarbya Fekrya ). A pre-post…

  20. Gap analysis survey: an aid in transitioning to standardized curricula for nuclear medicine technology.

    PubMed

    Bires, Angela Macci; Mason, Donna L; Gilmore, David; Pietrzyk, Carly

    2012-09-01

    This article discusses the process by which the Society of Nuclear Medicine Technology Section (SNMTS) is assisting educators as they transition to comply with the fourth edition of the Curriculum Guide for Educational Programs in Nuclear Medicine Technology. An electronic survey was sent to a list of nuclear medicine technology programs compiled by the educational division of the SNMTS. The collected data included committee member demographics, goals and objectives, conference call minutes, consultation discussions, transition examples, 4- and 2-y program curricula, and certificate program curricula. There were 56 responses to the survey. All respondents were program directors, with 3 respondents having more than one type of program, for a total of 59 programs. Of these, 19 (33.93%) were baccalaureate, 19 (28.57%) associate, and 21 (37.5%) certificate. Forty-eight respondents (85.71%) had accreditation through the Joint Review Commission on Educational Programs in Nuclear Medicine Technology, 6 (10.71%) had regional accreditation, and 2 (3.57%) were accredited through other entities. Thirteen categories of required general education courses were identified, and the existing program curricula of 9 (69.2%) courses were more than 50% compliant with the fourth edition Curriculum Guide. The fact that no measurable gap could be found within the didactic professional content across programs was due to the lack of a degree requirement and content standardization within the profession. The data indicated that the participating programs offer a minimum of 1-15 contact hours in emerging technology modalities. The required clinical hours ranged from 765 to 1,920 for degree or certificate completion. The average number of clinical hours required for all programs was 1,331.69. Standardization of the number and types of courses is needed both for current baccalaureate programs and for clinical education. This standardization will guide programs in transitioning from a certificate or associate level to the baccalaureate level. The greatest obstacle is in expanding curricula to meet the recommendations of the fourth edition Curriculum Guide. Such expansion to entry-level competency may be met by incorporating hybrid imaging courses, secondary-level courses, and equivalency courses on the basic sciences and emerging technologies.

  1. Outcome of Dupuytren Contractures After Collagenase Clostridium Histolyticum Injection: A Single-institution Experience.

    PubMed

    Hwee, Yin Kan; Park, Daniel; Vinas, Marisa; Litts, Christopher; Friedman, David

    2017-08-01

    Collagenase clostridium histolyticum (CCH) injection is an alternative to surgery for patients with Dupuytren disease (DD) of the metacarpophalangeal (MCP) and proximal interphalangeal (PIP) joints. The success of surgical and nonsurgical treatment modalities for DD is reported to vary widely between 25% and 80% (J Bone Joint Surg Am. 1985;67:1439-1443; Plast Reconstr Surg. 2007;120:44e-54e; J Bone Joint Surg Am. 2007;89:189-198; J Hand Surg Am. 2011:36:936-942; J Hand Surg Am. 1990;15:755-761; J Hand Surg Br. 1996;21:797-800; J Bone Joint Surg Br. 2000;82:90-94; Plast Reconstr Surg. 2005;115:802-810; Ann Plast Surg. 2006;57:13-17). This study presents the outcomes of patients with DD contractures treated with CCH injections at a single institution. An institutional review board-approved retrospective study was conducted of patients with DD of the hand treated with CCH injections in a single institution from February 2010 to April 2015. All patients received the recommended dose of 0.58 mg of CCH and returned for joint manipulation the following day. Data for follow-up at 7 and 30 days postoperatively and up to 5 years for patients who returned seeking further therapy for recurrent symptoms were reviewed. One hundred thirteen patients with a total of 146 affected joints (72 MCP; 74 PIP) were treated with CCH injections (95 males; 18 females; age, 40-92 y). Successful CCH therapy occurred in 75% of injected joints (109/146 joints; 59 MCP; 50 PIP), as defined by less than 5 degrees of contracture after treatment. Twenty-three percent of treated joints had partial correction (34/146 joints; 13 MCP; 21 PIP), as defined by between 5 and 30 degrees of residual contracture after treatment. Three patients (2%) had a failure of treatment, as defined by unchanged or worsened contracture from pretreatment baseline measurements. Fifteen patients (13%) returned to the clinic seeking additional therapy for recurrent joint contracture symptoms in 17 joints over a span of 1.5 months to 4 years after initial successful or partially successful treatment (17/143, 12%; 5 MCP; 12 PIP). Recurrence was defined as patients who sought treatment for a return of symptoms or greater than 20 degrees contracture in the setting of a palpable cord after initial full or partial contracture correction. Our 5-year outcome of CCH injections for DD contractures revealed full correction in 75% and partial correction in 23% of treated joints, with failure of treatment seen in only 2% of patients. Thirteen percent of the patients returned for additional treatment because of symptoms resulting from contracture recurrence in 12% of initially corrected or partially corrected joints. These positive outcomes are comparable with current surgical treatment modalities (J Hand Surg Am. 1990;15:755-761; J Bone Joint Surg Am. 1962;44B:602-613; J Clin Epidemiol. 2000;53:291-296). The use of CCH injections is an important nonsurgical treatment alternative for DD contractures of the MCP and PIP joints.

  2. Mandibular kinematics represented by a non-orthogonal floating axis joint coordinate system.

    PubMed

    Leader, Joseph K; Boston, J Robert; Debski, Richard E; Rudy, Thomas E

    2003-02-01

    There are many methods used to represent joint kinematics (e.g., roll, pitch, and yaw angles; instantaneous center of rotation; kinematic center; helical axis). Often in biomechanics internal landmarks are inferred from external landmarks. This study represents mandibular kinematics using a non-orthogonal floating axis joint coordinate system based on 3-D geometric models with parameters that are "clinician friendly" and mathematically rigorous. Kinematics data for two controls were acquired from passive fiducial markers attached to a custom dental clutch. The geometric models were constructed from MRI data. The superior point along the arc of the long axis of the condyle was used to define the coordinate axes. The kinematic data and geometric models were registered through fiducial markers visible during both protocols. The mean absolute maxima across the subjects for sagittal rotation, coronal rotation, axial rotation, medial-lateral translation, anterior-posterior translation, and inferior-superior translation were 34.10 degrees, 1.82 degrees, 1.14 degrees, 2.31, 21.07, and 6.95 mm, respectively. All the parameters, except for one subject's axial rotation, were reproducible across two motion recording sessions. There was a linear correlation between sagittal rotation and translation, the dominant motion plane, with approximately 1.5 degrees of rotation per millimeter of translation. The novel approach of combining the floating axis system with geometric models succinctly described mandibular kinematics with reproducible and clinician friendly parameters.

  3. Rehabilitation after cell transplantation for cartilage defects.

    PubMed

    Deszczynski, J; Slynarski, K

    2006-01-01

    Rehabilitation is a key element of successful treatment of cartilage defects with cell transplantation. The process of graft maturation takes approximately 18 months and cannot be accelerated, but requires carefully introduced steps leading to early recovery of joint function. Rehabilitation starts at 8 hours after surgery with the continuous passive motion (CPM) exercises and physiotherapy. For the first 6 weeks, patients continue with CPM in the range of 0 degrees to 45 degrees for femoral and tibial defects and 0 degrees to 30 degrees for patellofemoral joint reconstruction. Isometric muscle training and scar manual therapy are introduced. Patients are allowed to weight-bear as tolerated from the second week after surgery. After this initial phase, from 6 to 8 weeks after surgery, rehabilitation is accelerated with increased load-bearing and progressive range of motion to full flexion. Usually patients are able to walk without crutches in this time. Proprioceptive training is introduced with the advance of pain-free full range of motion and no discomfort with full weight-bearing. At 6 months after surgery, most patients recover joint function, making it possible for them to return to daily living activities. However, they need to continue with muscle, proprioceptive, and sports-specific rehabilitation exercises. The rehabilitation process is complicated, requiring close cooperation between the patient and surgeon-physiotherapist team to understand the symptoms and address them in a timely fashion.

  4. Ehlers-Danlos Syndrome in Orthopaedics

    PubMed Central

    Shirley, Eric D.; DeMaio, Marlene; Bodurtha, Joanne

    2012-01-01

    Ehlers-Danlos syndrome is a heterogeneous connective tissue condition characterized by varying degrees of skin hyperextensibility, joint hypermobility, and vascular fragility. Joint dislocations, musculoskeletal pain, atrophic scars, easy bleeding, vessel/viscera rupture, severe scoliosis, and obstetric complications may occur. These manifestations are secondary to abnormal collagen, with specific molecular defects in types I, III, and V collagen; they may also be related to tenascin-X, which has been identified in some patients. Ehlers-Danlos syndrome has been classified into 6 types, with variable degrees of joint instability, skin hyperextensibility, wound healing difficulty, and vascular fragility. Diagnosis begins with recognition of the signs and symptoms of global hypermobility and referring appropriate patients for genetic consultation. It is important to accurately identify patients with Ehlers-Danlos syndrome to initiate appropriate musculoskeletal treatment, optimize anesthetic and postoperative management, perform appropriate vascular screening, and help families address their concerns with other families and advocacy groups. PMID:23016112

  5. Low temperature joining of ceramic composites

    DOEpatents

    Barton, Thomas J.; Anderson, Iver E.; Ijadi-Maghsoodi, Sina; Nosrati, Mohammad; Unal, Ozer

    1999-01-12

    A method of joining similar or dissimilar ceramic and ceramic composite materials, such as SiC continuous fiber ceramic composites, at relatively low joining temperatures uses a solventless, three component bonding agent effective to promote mechanical bond toughness and elevated temperature strength to operating temperatures of approximately 1200 degrees C. The bonding agent comprises a preceramic precursor, an aluminum bearing powder, such as aluminum alloy powder, and mixtures of aluminum metal or alloy powders with another powder, and and boron powder in selected proportions. The bonding agent is disposed as an interlayer between similar or dissimilar ceramic or cermaic composite materials to be joined and is heated in ambient air or inert atmosphere to a temperature not exceeding about 1200 degrees C. to form a strong and tough bond joint between the materials. The bond joint produced is characterized by a composite joint microstructure having relatively soft, compliant aluminum bearing particulate regions dispersed in a ceramic matrix.

  6. Low temperature joining of ceramic composites

    DOEpatents

    Barton, Thomas J.; Anderson, Iver E.; Ijadi-Maghsoodi, Sina; Nosrati, Mohammad; Unal, Ozer

    1999-07-13

    A method of joining similar or dissimilar ceramic and ceramic composite materials, such as SiC continuous fiber ceramic composites, at relatively low joining temperatures uses a solventless, three component bonding agent effective to promote mechanical bond toughness and elevated temperature strength to operating temperatures of approximately 1200 degrees C. The bonding agent comprises a preceramic precursor, an aluminum bearing powder, such as aluminum alloy powder, and mixtures of aluminum metal or alloy powders with another powder, and and boron powder in selected proportions. The bonding agent is disposed as an interlayer between similar or dissimilar ceramic or ceramic composite materials to be joined and is heated in ambient air or inert atmosphere to a temperature not exceeding about 1200 degrees C. to form a strong and tough bond joint between the materials. The bond joint produced is characterized by a composite joint microstructure having relatively soft, compliant aluminum bearing particulate regions dispersed in a ceramic matrix.

  7. Low temperature joining of ceramic composites

    DOEpatents

    Barton, Thomas J.; Anderson, Iver E.; Ijadi-Maghsoodi, Sina; Nosrati, Mohammad; Unal, Ozer

    2001-04-10

    A method of joining similar or dissimilar ceramic and ceramic composite materials, such as SiC continuous fiber ceramic composites, at relatively low joining temperatures uses a solventless, three component bonding agent effective to promote mechanical bond toughness and elevated temperature strength to operating temperatures of approximately 1200 degrees C. The bonding agent comprises a preceramic precursor, an aluminum bearing powder, such as aluminum alloy powder, and mixtures of aluminum metal or alloy powders with another powder, and and boron powder in selected proportions. The bonding agent is disposed as an interlayer between similar or dissimilar ceramic or cermaic composite materials to be joined and is heated in ambient air or inert atmosphere to a temperature not exceeding about 1200 degrees C. to form a strong and tough bond joint between the materials. The bond joint produced is characterized by a composite joint microstructure having relatively soft, compliant aluminum bearing particulate regions dispersed in a ceramic matrix.

  8. Low temperature joining of ceramic composites

    DOEpatents

    Barton, T.J.; Anderson, I.E.; Ijadi-Maghsoodi, S.; Nosrati, M.; Unal, O.

    1999-01-12

    A method of joining similar or dissimilar ceramic and ceramic composite materials, such as SiC continuous fiber ceramic composites, at relatively low joining temperatures uses a solventless, three component bonding agent effective to promote mechanical bond toughness and elevated temperature strength to operating temperatures of approximately 1200 degrees C. The bonding agent comprises a preceramic precursor, an aluminum bearing powder, such as aluminum alloy powder, and mixtures of aluminum metal or alloy powders with another powder, and boron powder in selected proportions. The bonding agent is disposed as an interlayer between similar or dissimilar ceramic or ceramic composite materials to be joined and is heated in ambient air or inert atmosphere to a temperature not exceeding about 1200 degrees C. to form a strong and tough bond joint between the materials. The bond joint produced is characterized by a composite joint microstructure having relatively soft, compliant aluminum bearing particulate regions dispersed in a ceramic matrix. 3 figs.

  9. 75 FR 13643 - ITS Joint Program Office; Intelligent Transportation Systems Program Advisory Committee; Notice...

    Federal Register 2010, 2011, 2012, 2013, 2014

    2010-03-22

    ... Advisory Committee; Notice of Meeting AGENCY: Research and Innovative Technology Administration, U.S... Plan; (5) Evolution of IntelliDrive\\SM\\; (6) ITS Strategic Research Plan, 2010-2014; (7) University... Technology Administration, ITS Joint Program Office, Attention: Stephen Glasscock, 1200 New Jersey Avenue, SE...

  10. Soils, surficial geology, and geomorphology of the Bear Creek Valley Low-Level Waste Disposal Development and Demonstration Program site

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lietzke, D.A.; Lee, S.Y.; Lambert, R.E.

    1988-04-01

    An intensive soil survey was conducted on the proposed Low-Level Waste Disposal Development and Demonstration Program site (LLWDDD) in Bear Creek Valley. Soils on the site were related to the underlying residuum and to the surficial colluvium and alluvium. Within any particular geologic formation, soils were subdivided based mostly on the degree of weathering, as reflected by saprolite weathering and morphologic features of the soils. Degree of weathering was related both to slope shape and gradient and to the joint-fracture system. Erosion classes were also used to make further subdivisions of any particular soil. Deep pits were dug in eachmore » of the major Conasauga Group formations (Pumpkin Valley, Rogersville, Maryville, and Nolichucky) for soil and saprolite characterization. Because of the widespread presence of alluvium and colluvium, which are potential sources of fill and final cover material, pits and trenches were dug to characterize the properties of these soils and to try to understand the past geomorphic history of the site. The results of the soil survey investigation indicated that the deeply weathered Pumpkin Valley residuum has good potential for the construction of tumuli or other types of belowground or aboveground burial of prepackaged compacted waste. 11 refs., 30 figs., 3 tabs.« less

  11. International Internships in Nuclear Safeguards and Security: Challenges and Successes

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Duncan, Cristen L.; Heinberg, Cynthia L.; Killinger, Mark H.

    2010-04-20

    All students in the Russian safeguards and security degree programs at the National Research Nuclear University MEPhI and Tomsk Polytechnic University, sponsored by the Material Protection, Control and Accounting (MPC&A) Education Project, take part in a domestic internship at a Russian enterprise or facility. In addition, a select few students are placed in an international internship. These internships provide students with a better view of how MPC&A and nonproliferation in general are addressed outside of Russia. The possibility of an international internship is a significant incentive for students to enroll in the safeguards and security degree programs. The U.S. membersmore » of the MPC&A Education Project team interview students who have been nominated by their professors. These students must have initiative and reasonable English skills. The project team and professors then select students to be tentatively placed in various international internships during the summer or fall of their final year of study. Final arrangements are then made with the host organizations. This paper describes the benefits of the joint United States/Russia cooperation for next-generation workforce development, some of the international internships that have been carried out, the benefits of these international internships, and lessons learned in implementing them.« less

  12. Kinematic rate control of simulated robot hand at or near wrist singularity

    NASA Technical Reports Server (NTRS)

    Barker, K.; Houck, J. A.; Carzoo, S. W.

    1985-01-01

    A robot hand should obey movement commands from an operator on a computer program as closely as possible. However, when two of the three rotational axes of the robot wrist are colinear, the wrist loses a degree of freedom, and the usual resolved rate equations (used to move the hand in response to an operator's inputs) are indeterminant. Furthermore, rate limiting occurs in close vicinity to this singularity. An analysis shows that rate limiting occurs not only in the vicinity of this singularity but also substantially away from it, even when the operator commands rotational rates of the robot hand that are only a small percentage of the operational joint rate limits. Therefore, joint angle rates are scaled when they exceed operational limits in a real time simulation of a robot arm. Simulation results show that a small dead band avoids the wrist singularity in the resolved rate equations but can introduce a high frequency oscillation close to the singularity. However, when a coordinated wrist movement is used in conjunction with the resolved rate equations, the high frequency oscillation disappears.

  13. Effects of variables upon pyrotechnically induced shock response spectra, part 2

    NASA Technical Reports Server (NTRS)

    Smith, James Lee

    1988-01-01

    Throughout the aerospace industry, large variations of 50 percent (6 dB) or more in shock response spectra (SRS) derived from pyrotechnic separation events continue to be reported from actual spaceflight data and from laboratory tests. As a result of these variations, NASA funded a research program for 1984 through 1986. The purpose of the 1984 through 1986 project was to analyze variations in pyrotechnically induced SRS and to determine if and to what degree manufacturing and assembly variables and tolerances, distance from the shock source, data acquisition instrumentation, and shock energy propagation affect the SRS. Sixty-four free-free boundary plate tests were performed. NASA funded an additional study for 1987 through 1988. This paper is a summary of the additional study. The purpose was to evaluate shock dissipation through various spacecraft structural joint types, to evaluate shock variation for various manufacturing and assembly variables on clamped boundary test plates, and to verify data correction techniques. Five clamped boundary plate tests investigated manufacturing and assembly variables and mass loading effects. Six free-free boundary plate tests investigated shock dissipation across spacecraft joint structures.

  14. Cupiennius salei: biomechanical properties of the tibia-metatarsus joint and its flexing muscles.

    PubMed

    Siebert, Tobias; Weihmann, Tom; Rode, Christian; Blickhan, Reinhard

    2010-02-01

    Hunting spiders are well adapted to fast locomotion. Space saving hydraulic leg extension enables leg segments, which consist almost soley of flexor muscles. As a result, the muscle cross sectional area is high despite slender legs. Considering these morphological features in context with the spider's segmented C-shaped legs, these specifics might influence the spider's muscle properties. Moreover, these properties have to be known for modeling of spider locomotion. Cupiennius salei (n = 5) were fixed in a metal frame allowing exclusive flexion of the tibia-metatarsus joint of the second leg (counted from anterior). Its flexing muscles were stimulated supramaximally using needle electrodes. Accounting for the joint geometry, the force-length and the force-velocity relationships were determined. The spider muscles produce 0.07 N cm maximum isometric moment (corresponding to 25 N/cm(2) maximum stress) at 160 degrees tibia-metatarsus joint angle. When overextended to the dorsal limit at approximately 200 degrees , the maximum isometric moments decrease to 72%, and, when flexed to the ventral hinge stop at 85 degrees , they drop to 11%. The force-velocity relation shows the typical hyperbolic shape. The mean maximum shortening velocity is 5.7 optimum muscle lengths per second and the mean curvature (a/F (iso)) of the Hill-function is 0.34. The spider muscle's properties which were determined are similar to those of other species acting as motors during locomotion (working range, curvature of Hill hyperbola, peak power at the preferred speeds), but they are relatively slow. In conjunction with the low mechanical advantage (muscle lever/load arm), the arrangement of three considerably actuated joints in series may nonetheless enable high locomotion velocities.

  15. Torso and Bowing Arm Three-Dimensional Joint Kinematics of Elite Cellists: Clinical and Pedagogical Implications for Practice.

    PubMed

    Hopper, Luke; Chan, Cliffton; Wijsman, Suzanne; Ackland, Timothy; Visentin, Peter; Alderson, Jacqueline

    2017-06-01

    Elite cello playing requires complex and refined motor control. Cellists are prone to right shoulder and thoracolumbar injuries. Research informing injury management of cellists and cello pedagogy is limited. The aims of this study were to quantify the torso, right shoulder, and elbow joint movement used by elite cellists while performing a fundamental playing task, a C major scale, under two volume conditions. An eight degrees of freedom upper limb biomechanical model was applied to 3D motion capture data of the torso, upper arm, and forearm for 31 cellists with a mean experience of 19.4 yrs (SD 9.1). Two-factor ANOVA compared the joint positions between the four cello strings and two volume conditions. Significant (p<0.05) effects were found for either the string and/or volume conditions across all torso, shoulder, and elbow joint degrees of freedom. The torso was consistently positioned in left rotation from 5.0° (SD 5.6) at the beginning of the scale, increasing to 16.3° (5.5) at its apogee. The greatest mean shoulder flexion, internal rotation, and abduction joint angles were observed when playing at the tip of the bow on the top string (A): 107.2° (11.6), 59.1° (7.1), and -76.9° (15.7), respectively, during loud playing. Elite cellists use specific movement patterns to achieve string crossings and volume regulation during fundamental playing tasks. Implications of the static left-rotated torso posture and high degrees of combined shoulder flexion and internal rotation can be used to inform clinical and pedagogical practices.

  16. Staged lengthening arthroplasty for pediatric osteosarcoma around the knee.

    PubMed

    Kong, Chang-Bae; Lee, Soo-Yong; Jeon, Dae-Geun

    2010-06-01

    Orthopaedic oncologists often must address leg-length discrepancy after resection of tumors in growing patients with osteosarcoma. There are various alternatives to address this problem. We describe a three-stage procedure: (1) temporary arthrodesis, (2) lengthening by Ilizarov apparatus, and (3) tumor prosthesis. We asked (1) to what extent are affected limbs actually lengthened; (2) how many of the patients who undergo a lengthening procedure eventually achieve joint arthroplasty; and (3) can the three-stage procedure give patients a functioning joint with equalization of limb length? We reviewed 56 patients (younger than 14 years) with osteosarcoma who had staged lengthening arthroplasty between 1991 and 2004. Thirty-five of the 56 patients (63%) underwent soft tissue lengthening, and of these 35, 28 (50% of the original group of 56) had implantation of a mobile joint. Three of the 28 prostheses were later removed owing to infection after arthroplasty. The overall average length gained was 7.8 cm (range, 4-14 cm), and 25 (71%) of the 35 patients had a mobile joint at final followup. The average Musculoskeletal Tumor Society functional score was 23.2 (range, 15-28) and limb-length discrepancy at final followup was 2.6 cm (range, 0-6.5 cm). Although most mobile joints had an acceptable ROM (average, 74.2 degrees ; range, 35 degrees -110 degrees ), extension lag was frequent. Our approach is one option for skeletally immature patients, especially in situations where an expandable prosthesis is not available. However, this technique requires multiple stages and would be inappropriate for patients who cannot accept prolonged functional deficit owing to a limited lifespan or other reasons. Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.

  17. Growing multiplex networks with arbitrary number of layers

    NASA Astrophysics Data System (ADS)

    Momeni, Naghmeh; Fotouhi, Babak

    2015-12-01

    This paper focuses on the problem of growing multiplex networks. Currently, the results on the joint degree distribution of growing multiplex networks present in the literature pertain to the case of two layers and are confined to the special case of homogeneous growth and are limited to the state state (that is, the limit of infinite size). In the present paper, we first obtain closed-form solutions for the joint degree distribution of heterogeneously growing multiplex networks with arbitrary number of layers in the steady state. Heterogeneous growth means that each incoming node establishes different numbers of links in different layers. We consider both uniform and preferential growth. We then extend the analysis of the uniform growth mechanism to arbitrary times. We obtain a closed-form solution for the time-dependent joint degree distribution of a growing multiplex network with arbitrary initial conditions. Throughout, theoretical findings are corroborated with Monte Carlo simulations. The results shed light on the effects of the initial network on the transient dynamics of growing multiplex networks and takes a step towards characterizing the temporal variations of the connectivity of growing multiplex networks, as well as predicting their future structural properties.

  18. [Forensic identification of floating shoulder injury].

    PubMed

    Li, Sheng-Ya; Huang, Si-Xing; Zhao, Xing-Bin; Zheng, Xiao-Hong; Zhu, Yi

    2006-10-15

    To discuss forensic identification of floating shoulder injury (FSI). To analyze fifteen cases of FSI which were accepted from Jan. 1993 to Jan. 2006, including 15 shoulder neck fracture, 13 clavide stem fracture and 2 distal end clavide fracture, the function of shoulder joint was evaluated six months after injure considering the following three aspects: result of forensic examination such as X-ray photograph, CT and MRI, the injurers' symptom, objective sign and joint function, shoulder joint territory, degree of pain and local muscle power. Basing on the curative effect standard of Herscovic, all cases were divided into good. Modest, worst, which included 2, 4, 9 cases respectively; referring the standard of GA35-92, GB18667-2002, all cases were divided into six, seven, eight, nine and ten degree, which included 2,9,2,1,1 cases respectively. As a special powerful injure, FSI always companied with concurrent and multiple injure, and characterized by missed, incorrect and delayed diagnosis and infelicitous treatment, which lead to the high frequency and degree of injure. To prevent missed and incorrect forensic identification, we should have a full realization of the particularity of FSI, and evaluate the function of shoulder all-sidely, objectively and synseticaly.

  19. Decline eccentric squats increases patellar tendon loading compared to standard eccentric squats.

    PubMed

    Kongsgaard, M; Aagaard, P; Roikjaer, S; Olsen, D; Jensen, M; Langberg, H; Magnusson, S P

    2006-08-01

    Recent studies have shown excellent clinical results using eccentric squat training on a 25 degrees decline board to treat patellar tendinopathy. It remains unknown why therapeutic management of patellar tendinopathy using decline eccentric squats offer superior clinical efficacy compared to standard horizontal eccentric squats. This study aimed to compare electromyography activity, patellar tendon strain and joint angle kinematics during standard and decline eccentric squats. Thirteen subjects performed unilateral eccentric squats on flat-and a 25 degrees decline surface. During the squats, electromyography activity was obtained in eight representative muscles. Also, ankle, knee and hip joint goniometry was obtained. Additionally, patellar tendon strain was measured in vivo using ultrasonography as subjects maintained a unilateral isometric 90 degrees knee angle squat position on either flat or 25 degrees decline surface. Patellar tendon strain was significantly greater (P<0.05) during the squat position on the decline surface compared to the standard surface. The stop angles of the ankle and hip joints were significantly smaller during the decline compared to the standard squats (P<0.001, P<0.05). Normalized mean electromyography amplitudes of the knee extensor muscles were significantly greater during the decline compared to the standard squats (P<0.05). Hamstring and calf muscle mean electromyography did not differ, respectively, between standard and decline squats. The use of a 25 degrees decline board increases the load and the strain of the patellar tendon during unilateral eccentric squats. This finding likely explains previous reports of superior clinical efficacy of decline eccentric squats in the rehabilitative management of patellar tendinopathy.

  20. Joint torques and joint reaction forces during squatting with a forward or backward inclined Smith machine.

    PubMed

    Biscarini, Andrea; Botti, Fabio M; Pettorossi, Vito E

    2013-02-01

    We developed a biomechanical model to determine the joint torques and loadings during squatting with a backward/forward-inclined Smith machine. The Smith squat allows a large variety of body positioning (trunk tilt, foot placement, combinations of joint angles) and easy control of weight distribution between forefoot and heel. These distinctive aspects of the exercise can be managed concurrently with the equipment inclination selected to unload specific joint structures while activating specific muscle groups. A backward (forward) equipment inclination decreases (increases) knee torque, and compressive tibiofemoral and patellofemoral forces, while enhances (depresses) hip and lumbosacral torques. For small knee flexion angles, the strain-force on the posterior cruciate ligament increases (decreases) with a backward (forward) equipment inclination, whereas for large knee flexion angles, this behavior is reversed. In the 0 to 60 degree range of knee flexion angles, loads on both cruciate ligaments may be simultaneously suppressed by a 30 degree backward equipment inclination and selecting, for each value of the knee angle, specific pairs of ankle and hip angles. The anterior cruciate ligament is safely maintained unloaded by squatting with backward equipment inclination and uniform/forward foot weight distribution. The conditions for the development of anterior cruciate ligament strain forces are clearly explained.

  1. Interjoint coupling effects on muscle contributions to endpoint force and acceleration in a musculoskeletal model of the cat hindlimb

    PubMed Central

    van Antwerp, Keith W.; Burkholder, Thomas J.

    2015-01-01

    The biomechanical principles underlying the organization of muscle activation patterns during standing balance are poorly understood. The goal of this study was to understand the influence of biomechanical inter-joint coupling on endpoint forces and accelerations induced by the activation of individual muscles during postural tasks. We calculated induced endpoint forces and accelerations of 31 muscles in a 7 degree-of-freedom, 3-dimensional model of the cat hindlimb. To test the effects of inter-joint coupling, we systematically immobilized the joints (excluded kinematic degrees-of-freedom) and evaluated how the endpoint force and acceleration directions changed for each muscle in seven different conditions. We hypothesized that altered inter-joint coupling due to joint immobilization of remote joints would substantially change the induced directions of endpoint force and acceleration of individual muscles. Our results show that for most muscles crossing the knee or the hip, joint immobilization altered the endpoint force or acceleration direction by more than 90° in the dorsal and sagittal planes. Induced endpoint forces were typically consistent with behaviorally-observed forces only when the ankle was immobilized. We then activated a proximal muscle simultaneous with an ankle torque of varying magnitude, which demonstrated that the resulting endpoint force or acceleration direction is modulated by the magnitude of the ankle torque. We argue that this simple manipulation can lend insight into the functional effects of co-activating muscles. We conclude that inter-joint coupling may be an essential biomechanical principle underlying the coordination of proximal and distal muscles to produce functional endpoint actions during motor tasks. PMID:17640652

  2. Mutual coordination strengthens the sense of joint agency in cooperative joint action.

    PubMed

    Bolt, Nicole K; Poncelet, Evan M; Schultz, Benjamin G; Loehr, Janeen D

    2016-11-01

    Philosophers have proposed that when people coordinate their actions with others they may experience a sense of joint agency, or shared control over actions and their effects. However, little empirical work has investigated the sense of joint agency. In the current study, pairs coordinated their actions to produce tone sequences and then rated their sense of joint agency on a scale ranging from shared to independent control. People felt more shared than independent control overall, confirming that people experience joint agency during joint action. Furthermore, people felt stronger joint agency when they (a) produced sequences that required mutual coordination compared to sequences in which only one partner had to coordinate with the other, (b) held the role of follower compared to leader, and (c) were better coordinated with their partner. Thus, the strength of joint agency is influenced by the degree to which people mutually coordinate with each other's actions. Copyright © 2016 Elsevier Inc. All rights reserved.

  3. Joint University Program for Air Transportation Research, 1987

    NASA Technical Reports Server (NTRS)

    Morrell, Frederick R. (Compiler)

    1989-01-01

    The research conducted during 1987 under the NASA/FAA sponsored Joint University Program for Air Transportation Research is summarized. The Joint University Program is a coordinated set of 3 grants sponsored by NASA-Langley and the FAA, one each with the MIT, Ohio Univ., and Princeton Univ. Completed works, status reports, and annotated bibliographies are presented for research topics, which include computer science, guidance and control theory and practice, aircraft performance, flight dynamics, and applied experimental psychology. An overview of the year's activities for each university is also presented.

  4. Tech-Prep/Associate Degree Program Guide: Tech Prep Associate Degree Program, Business Administration Associate Degree Program, Office Administration Associate Degree Program, Allied Health Associate Degree Program.

    ERIC Educational Resources Information Center

    Marmaras, Judy; Neri, Pat

    The Tech-Prep Associate Degree Program (TPAD) at the Community College of Rhode Island (CCRI) in Warwick, is a high school/community college partnership providing high school students with an alternative program of study focused on goal setting, basic academic skills development, and the skills needed to pursue a career in a technical, business or…

  5. Bridging the Technology Valley of Death in Joint Medical Development

    DTIC Science & Technology

    2015-11-01

    Force lieutenant colonel, is the Air Force Medical Support Agency Advanced Development Liaison Field Engineer in Falls Church, Virginia. Prusaczyk is...Awareness, communication and coordination may be mini - mal among Service S&T and AD programs. Joint Transition Planning Process A Joint Transition...Human Proof of Phase III NDA/BLA ling Approval, Launch Concept*** Launch Review Program Initiation Materiel Technology Engineering & Production

  6. UTeach: Secondary Teacher Preparation in Science and Mathematics at the University of Texas at Austin

    NASA Astrophysics Data System (ADS)

    Marder, Michael

    2006-03-01

    The UTeach Program is a joint effort of the College of Natural Sciences, the College of Education and the Austin Independent School District to recruit, prepare and support math and science teachers for the State of Texas. UTeach uses early and on-going field experiences to capture the imagination of preservice teachers and provide a foundation for more advanced pedagogical courses. With over 400 students enrolled and over 80 graduates per year, UTeach is one of the largest programs producing secondary science, mathematics, and computer science teachers in the nation. Most UTeach students are undergraduates, but around 10% are people of many ages with strong backgrounds in mathematics or science who have decided to enter teaching. Hallmarks include: *Four-year degree plans that enable undergraduates to obtain certification at no cost in time or money. *Active recruitment and support including tuition reimbursement, paid internships, personal advising, and guidance by master teachers. *Emphasis on preparing teachers who will be knowledgeable of their discipline, experienced with involving students in scientific inquiry, and practiced in employing new technologies to enhance student learning. *A revised, streamlined professional education sequence drawing on research on learning, standards-based curricula, multiple forms of assessment, and proven strategies for achieving equity and integrating technology into math and science education. *Program flexibility with multiple entry points (from freshman to post baccalaureate), integrated degree plans, and proficiency-based assessment, including the development of individual teaching portfolios. For more information on UTeach, see http://uteach.utexas.edu

  7. 76 FR 21789 - ITS Joint Program Office; Vehicle to Infrastructure Core System Concept of Operations; Notice of...

    Federal Register 2010, 2011, 2012, 2013, 2014

    2011-04-18

    ... DEPARTMENT OF TRANSPORTATION ITS Joint Program Office; Vehicle to Infrastructure Core System... Program Office (ITS JPO) will host a free public meeting to discuss the Vehicle to Infrastructure (V2I... to work originally performed under the Vehicle Infrastructure Integration Proof of Concept (VII POC...

  8. Evaluation of an interprofessional education program for advanced practice nursing and dental students: The oral-systemic health connection.

    PubMed

    Nash, Whitney A; Hall, Lynne A; Lee Ridner, S; Hayden, Dedra; Mayfield, Theresa; Firriolo, John; Hupp, Wendy; Weathers, Chandra; Crawford, Timothy N

    2018-07-01

    In response to the growing body of evidence supporting the need for expanded interprofessional education among health professions, an interprofessional education program, based on the Interprofessional Education Collaborative Core Competencies, was piloted with nurse practitioner and dental students. The purpose of this pilot study was to evaluate a technology enhanced interprofessional education program focused on the oral-systemic health connection for nurse practitioner and dental students. A two-group comparative study using cross-sectional data and a quasi-experimental one-group pre-test/post-test design were used to evaluate students' knowledge of IPE core competencies, attitudes toward interprofessional education and interdisciplinary teamwork, and self-efficacy in functioning as a member of an interdisciplinary team. This program was implemented with master of science in nursing students pursuing a primary care nurse practitioner (NP) degree and dental students at a large urban academic health sciences center. Cohort 1 (N = 75) consisted of NP (n = 34) and dental students (n = 41) at the end of their degree program who participated in a one-time survey. Cohort 2 (N = 116) was comprised of second-year NP students (n = 22) and first-year dental students (n = 94) who participated in the IPE program. Students participated in a multi-faceted educational program consisting of technology- enhanced delivery as well as interactive exercises in the joint health assessment course. Data were collected prior to the initiation and at the conclusion of the program. Nurse practitioner and dental students who participated in the program had better self-efficacy in functioning as a member of an interdisciplinary team than graduating students who did not participate. Students from both nursing and dentistry who participated in the program had significantly improved self-efficacy in functioning in interprofessional teams from pre- to post-test. An interprofessional education program can be a valuable addition to the health professions curriculum of nurse practitioner and dental students. Care must be taken to address logistical issues when working with students in different academic programs. Copyright © 2018. Published by Elsevier Ltd.

  9. Joint Venture Arrangement for RN to BSN: A Model of Synergy between Academia and Service.

    ERIC Educational Resources Information Center

    Bargagliotti, L. Antoinette; And Others

    1991-01-01

    Joint venture among educational and practice institutions is well on its way toward becoming the norm in nursing education and practice. Kaiser Permanente and the University of San Francisco School of Nursing offer a venture that allows registered nurses to pursue a bachelor of science in nursing degree. (JOW)

  10. A self-aligning knee joint for walking assistance devices.

    PubMed

    Byungjune Choi; Younbaek Lee; Jeonghun Kim; Minhyung Lee; Jongwon Lee; Se-Gon Roh; Hyundo Choi; Yong-Jae Kim; Jung-Yun Choi

    2016-08-01

    This paper presents a novel self-aligning knee mechanism for walking assistance devices for the elderly to provide physical gait assistance. Self-aligning knee joints can assist in flexion/extension motions of the knee joint and compensate the knee's transitional movements in the sagittal plane. In order to compensate the center of rotation, which moves with the flexion/extension motion of the human knee joint, a self-aligning knee joint is proposed that adds redundant degrees of freedom (i.e., 2-DoF) to the 1-DoF revolute joint. The key idea of the proposed mechanism is to decouple joint rotations and translations for use in lower-extremity wearable devices. This paper describes the mechanical design of this self-aligning knee mechanism and its implementation on a wearable robot and in preliminary experiments. The performance of the proposed mechanism is verified by simulations and experiments.

  11. All joint moments significantly contribute to trunk angular acceleration

    PubMed Central

    Nott, Cameron R.; Zajac, Felix E.; Neptune, Richard R.; Kautz, Steven A.

    2010-01-01

    Computationally advanced biomechanical analyses of gait demonstrate the often counter intuitive roles of joint moments on various aspects of gait such as propulsion, swing initiation, and balance. Each joint moment can produce linear and angular acceleration of all body segments (including those on which the moment does not directly act) due to the dynamic coupling inherent in the interconnected musculoskeletal system. This study presents the quantitative relationships between individual joint moments and trunk control with respect to balance during gait to show that the ankle, knee, and hip joint moments all affect the angular acceleration of the trunk. We show that trunk angular acceleration is affected by all the joints in the leg with varying degrees of dependence during the gait cycle. Furthermore, it is shown that inter-planar coupling exists and a two dimensional analysis of trunk balance neglects important out-of-plane joint moments that affect trunk angular acceleration. PMID:20646711

  12. Bonded composite to metal scarf joint performance in an aircraft landing gear drag strut. [for Boeing 747 aircraft

    NASA Technical Reports Server (NTRS)

    Howell, W. E.

    1974-01-01

    The structural performance of a boron-epoxy reinforced titanium drag strut, which contains a bonded scarf joint and was designed to the criteria of the Boeing 747 transport, was evaluated. An experimental and analytical investigation was conducted. The strut was exposed to two lifetimes of spectrum loading and was statically loaded to the tensile and compressive design ultimate loads. Throughout the test program no evidence of any damage in the drag strut was detected by strain gage measurements, ultrasonic inspection, or visual observation. An analytical study of the bonded joint was made using the NASA structural analysis computer program NASTRAN. A comparison of the strains predicted by the NASTRAN computer program with the experimentally determined values shows excellent agreement. The NASTRAN computer program is a viable tool for studying, in detail, the stresses and strains induced in a bonded joint.

  13. A kinematic analysis of the modified flight telerobotic servicer manipulator system

    NASA Technical Reports Server (NTRS)

    Crane, Carl; Carnahan, Tim; Duffy, Joseph

    1992-01-01

    A reverse kinematic analysis is presented of a six-DOF subchain of a modified seven-DOF flight telerobotic servicer manipulator system. The six-DOF subchain is designated as a TR-RT chain, which describes the sequence of manipulator joints beginning with the first grounded hook joint (universal joint) T, where the sequence R-R designates a pair of revolute joints with parallel axes. At the outset, it had been thought that the reverse kinematic analysis would be similar to a TTT manipulator previously analyzed, in which the third and fourth joints intersected at a finite point. However, this is shown not the case, and a 16th-degree tan-half-angle polynomial is derived for the TR-RT manipulator.

  14. Development of an Underactuated 2-DOF Wrist Joint using McKibben PAMs

    NASA Astrophysics Data System (ADS)

    Rajagopal, S. P.; Jain, S.; Ramasubramanian, S. N.; Johnson, B. V.; Dwivedy, S. K.

    2014-10-01

    In this work, model of an underactuated 2-DOF wrist joint with pneumatically actuated muscles is proposed. For the joint, McKibben-type artificial muscles are used in parallel configuration for the actuation. For each Degree of Freedom (DOF) one agonist-antagonist pair arrangement is usually used with a pulley mechanism. A mathematical model of wrist joint is derived using conventional forward kinematic analysis. The static model relating pressure in the muscle with the orientation of the wrist joint is obtained by combining the experimental data and mathematical model. Regulation of pressure can be achieved by pulse width modulation control of on/off solenoid valves. A set of free vibration experiments are done for the dynamic identification of the muscle characteristics.

  15. Exoskeleton for gait rehabilitation of children: Conceptual design.

    PubMed

    Cornejo, Jorge L; Santana, Jesus F; Salinas, Sergio A

    2017-07-01

    This paper presents the conceptual design of an exoskeleton for gait rehabilitation of children. This system has electronics, mechanicals and software sections, which are implemented and tested using a mannequin of a child. The prototype uses servomotors to move robotic joints that are attached to simulated patient's legs. The design has 4 DOF (degrees of freedom) two for hip joints and other two for knee joints, in the sagittal plane. A microcontroller measures sensor signals, controls motors and exchanges data with a computer. The user interacts with a graphical interface to configure, control and monitor the exoskeleton activities. The laboratory tests show soften movements in joint angle tracking.

  16. [Non vascularized toe phalangeal transfers for symbrachydactyly. Active range of motion without joint reconstruction].

    PubMed

    Leca, J-B; Auquit Auckbur, I; Bachy, B; Milliez, P-Y

    2008-12-01

    Symbrachydactyly is a rare congenital malformation of the hand and its treatment is controversial. Non vascularized toe phalangeal transfers have been used for management of short digits for three children. Six phalanges have been harvested complete with their periosteum. No joint reconstruction has been performed and all children have undergone surgery at a young age. Four of six digits involved have an active range of motion (range 30 to 105 degrees ). All authors who have reported active range of motion of toe phalangeal transfers have performed joint reconstruction. Here, we report obtaining active range of motion of phalangeal transfers without necessity of joint reconstruction.

  17. High-strength laser welding of aluminum-lithium scandium-doped alloys

    NASA Astrophysics Data System (ADS)

    Malikov, A. G.; Ivanova, M. Yu.

    2016-11-01

    The work presents the experimental investigation of laser welding of an aluminum alloy (system Al-Mg-Li) and aluminum alloy (system Al-Cu-Li) doped with Sc. The influence of nano-structuring of the surface layer welded joint by cold plastic deformation on the strength properties of the welded joint is determined. It is founded that, regarding the deformation degree over the thickness, the varying value of the welded joint strength is different for these aluminum alloys. The strength of the plastically deformed welded joint, aluminum alloys of the Al-Mg-Li and Al-Cu-Li systems reached 0.95 and 0.6 of the base alloy strength, respectively.

  18. 40 CFR 80.1461 - Who is liable for violations under the RFS program?

    Code of Federal Regulations, 2010 CFR

    2010-07-01

    ... that is committed by any of its subsidiaries. (d) Joint venture liability. Each partner to a joint venture is jointly and severally liable for any violation of this subpart that is committed by the joint venture operation. ...

  19. Joint Enrollment Report, 2014

    ERIC Educational Resources Information Center

    Iowa Department of Education, 2014

    2014-01-01

    The Iowa Department of Education collects information on joint enrollment in Iowa's 15 community colleges. Jointly enrolled students are high school students enrolled in community college credit coursework. Most jointly enrolled students enroll through Senior Year Plus (SYP) programs such as Postsecondary Enrollment Options (PSEO) and concurrent…

  20. Joint Enrollment Report, 2015

    ERIC Educational Resources Information Center

    Iowa Department of Education, 2015

    2015-01-01

    The Iowa Department of Education collects information on joint enrollment from Iowa's 15 community colleges. Jointly enrolled students are high school students enrolled in community college credit coursework. Most jointly enrolled students enroll through Senior Year Plus programs such as Postsecondary Enrollment Options (PSEO) and concurrent…

  1. Skin movement artefact assessment and compensation in the estimation of knee-joint kinematics.

    PubMed

    Lucchetti, L; Cappozzo, A; Cappello, A; Della Croce, U

    1998-11-01

    In three dimensional (3-D) human movement analysis using close-range photogrammetry, surface marker clusters deform and rigidly move relative to the underlying bone. This introduces an important artefact (skin movement artefact) which propagates to bone position and orientation and joint kinematics estimates. This occurs to the extent that those joint attitude components that undergo small variations result in totally unreliable values. This paper presents an experimental and analytical procedure, to be included in a subject-specific movement analysis protocol, which allows for the assessment of skin movement artefacts and, based on this knowledge, for their compensation. The effectiveness of this procedure was verified with reference to knee-joint kinematics and to the artefacts caused by the hip movements on markers located on the thigh surface. Quantitative validation was achieved through experimental paradigms whereby prior reliable information on the target joint kinematics was available. When position and orientation of bones were determined during the execution of a motor task, using a least-squares optimal estimator, but the rigid artefactual marker cluster movement was not dealt with, then knee joint translations and rotations were affected by root mean square errors (r.m.s.) up to 14 mm and 6 degrees, respectively. When the rigid artefactual movement was also compensated for, then r.m.s errors were reduced to less than 4 mm and 3 degrees, respectively. In addition, errors originally strongly correlated with hip rotations, after compensation, lost this correlation.

  2. Comparative assessment of knee joint models used in multi-body kinematics optimisation for soft tissue artefact compensation.

    PubMed

    Richard, Vincent; Cappozzo, Aurelio; Dumas, Raphaël

    2017-09-06

    Estimating joint kinematics from skin-marker trajectories recorded using stereophotogrammetry is complicated by soft tissue artefact (STA), an inexorable source of error. One solution is to use a bone pose estimator based on multi-body kinematics optimisation (MKO) embedding joint constraints to compensate for STA. However, there is some debate over the effectiveness of this method. The present study aimed to quantitatively assess the degree of agreement between reference (i.e., artefact-free) knee joint kinematics and the same kinematics estimated using MKO embedding six different knee joint models. The following motor tasks were assessed: level walking, hopping, cutting, running, sit-to-stand, and step-up. Reference knee kinematics was taken from pin-marker or biplane fluoroscopic data acquired concurrently with skin-marker data, made available by the respective authors. For each motor task, Bland-Altman analysis revealed that the performance of MKO varied according to the joint model used, with a wide discrepancy in results across degrees of freedom (DoFs), models and motor tasks (with a bias between -10.2° and 13.2° and between -10.2mm and 7.2mm, and with a confidence interval up to ±14.8° and ±11.1mm, for rotation and displacement, respectively). It can be concluded that, while MKO might occasionally improve kinematics estimation, as implemented to date it does not represent a reliable solution to the STA issue. Copyright © 2017 Elsevier Ltd. All rights reserved.

  3. Influence of bi- and tri-compartmental knee arthroplasty on the kinematics of the knee joint.

    PubMed

    Wünschel, Markus; Lo, Jiahsuan; Dilger, Torsten; Wülker, Nikolaus; Müller, Otto

    2011-01-27

    The cruciate ligaments are important stabilizers of the knee joint and determine joint kinematics in the natural knee and after cruciate retaining arthroplasty.No in vitro data is available to biomechanically evaluate the ability of the anterior cruciate ligament (ACL) to maintain knee joint kinematics after bicruciate-retaining bi-compartmental knee arthroplasty (BKA).Therefore, the objective of the current study was to investigate the kinematics of the natural knee joint, before and after installing bicruciate-retaining BKA and posterior cruciate retaining total knee arthroplasty. Specifically, we incorporated a dynamic knee simulator to simulate weight-bearing flexions on cadaveric knee specimen before and after surgical manipulations. In this cadaveric study we investigated rotational and translational tibiofemoral kinematics during simulated weight-bearing flexions of the intact knee, after bi-compartmental knee arthroplasty (BKA+), after resecting the ACL in BKA (BKA-), and after posterior cruciate retaining total knee arthroplasty (TKA). Rotation of BKA+ is closest to the intact knee joint, whereas TKA shows significant differences from 30 to 90 degree of flexion. Within the tested flexion range (15 to 90 degree of flexion), there was no significant difference in the anterior-posterior translation among intact, BKA+, and TKA knees. Resecting the ACL in BKA leads to a significant anterior tibial translation. BKA with intact cruciate ligaments resembles rotation and translation of the natural knee during a simulated weight-bearing flexion. It is a suitable treatment option for medial and patellofemoral osteoarthritis with advantages in rotational characteristics compared to TKA.

  4. Does increased femoral antetorsion predispose to cartilage lesions of the patellofemoral joint?

    PubMed

    Oppermann, Johannes; Bredow, Jan; Wissusek, Boris; Spies, Christian Karl; Boese, Christoph Kolja; Chang, Shi-Min; Eysel, Peer; Dargel, Jens

    2017-09-01

    The purpose of this study was to investigate whether there was a relationship between femoral neck antetorsion and the presence and pattern of osteoarthritis of the patellofemoral joint. It was hypothesized that an increased femoral neck antetorsion (1) correlates with osteoarthritic changes of the lateral facet of the patellofemoral joint and (2) correlates with an increased lateral trochlear height and a decreased sulcus angle. Seventy-eight formalin-embedded cadaveric lower extremities from thirty-nine subjects with a median age of 74 years (range 60-88) were used. Surrounding soft tissues of the lower limb were removed. The femoral neck antetorsion was measured and referenced to the transepicondylar axis and the posterior condylar line. The height of the medial and lateral facet of the trochlea and the sulcus angle was measured. The location and the degree of patellofemoral cartilage degeneration were recorded. A Pearson's correlation analysis was performed to correlate the femoral neck antetorsion with the measured knee parameters. No significant correlation could be found between the femoral antetorsion and cartilage degeneration of the lateral patellofemoral joint (n.s.), the height of the lateral trochlea (n.s.) and the sulcus angle (n.s.). This study could not document that the femoral neck antetorsion and subsequent internal rotation of the distal femur correlated with the degree of degeneration of the lateral facet of the patellofemoral joint. Clinically, femoral internal rotation may play a minor role in the development of lateral patellofemoral joint degeneration.

  5. [Repair and reconstruction for severe fracture and dislocation of ankle joint].

    PubMed

    Yin, Qingwei; Jiang, Yi; Xiao, Lianping; Li, Xiaodong; Fu, Jiaxin; Tian, Yonggang; Han, Liqiang; Liu, Zhi

    2008-06-01

    To summarize the technique and effect of the therapy for severe fracture and dislocation of ankle joint by operation. From March 2003 to February 2006, 76 cases were treated with primary open restoration and internal fixation for dislocated ankle joint fracture, with 47 males and 29 females, with the average age of 36.4 years (ranging from 18 years to 65 years). According to AO criterion, these fresh fractures were classified into 13 cases for type C3-1, 45 cases for type C3-2 and 18 cases for type C3-3. Based on the Gustilo-Anderson standard, 23 open fractures were classified into 17 cases for type II and 6 cases for type III A. The operation was delayed from 1 hours to 24 hours after the injury. All incisions healed at the first stage except 4 cases which delayed union because of simple infection by revision with ointment. A total of 72 cases were followed up, with the average time of 18.5 months (from 12 months to 35 months). The time of bone union was from 12 weeks to 24 weeks. The screws of fixation for lower tibia-fibula joint were found to be ruptured in 2 cases when further consultation was performed in the 16th and 20th week after the operation, respectively, and were broken within 1 year after the operation. These screws were taken out 12 weeks postoperative in 28 cases, while the whole internal fixations of the rest cases were taken out 1 year after the operation. The postoperative function of malleolus extended from 21.7 degrees to 26.8 degrees and flection from 38.5 degrees to 44.7 degrees. Assessed by the American Orthopaedic Foot and Ankle Society Clinical Rating Scales, 23 cases were excellent, 36 good, 13 fair, and the choiceness rate reached 81.94%. These procedures, together with reduction by twist after hospital, open and internal fixation in time, and parenchyma managed with internal fixation, are important to attain satisfactory effect for the treatment of severe fracture and dislocation of ankle joint.

  6. Assessment of the Joint Food Science Curriculum of Washington State University and the University of Idaho by Graduates and Their Employers

    ERIC Educational Resources Information Center

    Clark, Stephanie; McCurdy, Alan; Roy, Sharon; Smith, Denise

    2006-01-01

    Thirty-two recent graduates from the joint food science program of Washington State Univ. (WSU) and The Univ. of Idaho (UI) and 12 of their employers participated in a survey study to assess food science program outcomes. The objective of this study was to assess the joint curriculum in its ability to prepare undergraduate students for critical…

  7. 7 CFR 3405.7 - Joint project proposals.

    Code of Federal Regulations, 2010 CFR

    2010-01-01

    ... Agriculture Regulations of the Department of Agriculture (Continued) COOPERATIVE STATE RESEARCH, EDUCATION, AND EXTENSION SERVICE, DEPARTMENT OF AGRICULTURE HIGHER EDUCATION CHALLENGE GRANTS PROGRAM Program... strengthening higher education in the food and agricultural sciences. The goals of such joint initiatives should...

  8. Strategic innovation between PhD and DNP programs: Collaboration, collegiality, and shared resources.

    PubMed

    Edwards, Joellen; Rayman, Kathleen; Diffenderfer, Sandra; Stidham, April

    2016-01-01

    At least 111 schools and colleges of nursing across the nation provide both PhD and DNP programs (AACN, 2014a). Collaboration between nurses with doctoral preparation as researchers (PhD) and practitioners (DNP) has been recommended as essential to further the profession; that collaboration can begin during the educational process. The purpose of this paper is to describe the development and implementation of successful DNP and PhD program collaboration, and to share the results of that collaboration in an educational setting. Faculty set strategic goals to maximize the effectiveness and efficiency of both new DNP and existing PhD programs. The goals were to promote collaboration and complementarity between the programs through careful capstone and dissertation differentiation, complementary residency activities, joint courses and inter-professional experiences; promote collegiality in a blended on-line learning environment through shared orientation and intensive on-campus sessions; and maximize resources in program delivery through a supportive organizational structure, equal access to technology support, and shared faculty responsibilities as appropriate to terminal degrees. Successes such as student and faculty accomplishments, and challenges such as managing class size and workload, are described. Collaboration, collegiality and the sharing of resources have strengthened and enriched both programs and contributed to the success of students, faculty. These innovative program strategies can provide a solid foundation for DNP and PhD collaboration. Copyright © 2016 Elsevier Inc. All rights reserved.

  9. Generating Long Scale-Length Plasma Jets Embedded in a Uniform, Multi-Tesla Magnetic-Field

    NASA Astrophysics Data System (ADS)

    Manuel, Mario; Kuranz, Carolyn; Rasmus, Alex; Klein, Sallee; Fein, Jeff; Belancourt, Patrick; Drake, R. P.; Pollock, Brad; Hazi, Andrew; Park, Jaebum; Williams, Jackson; Chen, Hui

    2013-10-01

    Collimated plasma jets emerge in many classes of astrophysical objects and are of great interest to explore in the laboratory. In many cases, these astrophysical jets exist within a background magnetic field where the magnetic pressure approaches the plasma pressure. Recent experiments performed at the Jupiter Laser Facility utilized a custom-designed solenoid to generate the multi-tesla fields necessary to achieve proper magnetization of the plasma. Time-gated interferometry, Schlieren imaging, and proton radiography were used to characterize jet evolution and collimation under varying degrees of magnetization. Experimental results will be presented and discussed. This work is funded by the NNSA-DS and SC-OFES Joint Program in High-Energy-Density Laboratory Plasmas, grant number DE-NA0001840, by the National Laser User Facility Program, grant number DE-NA0000850, by the Predictive Sciences Academic Alliances Program in NNSA-ASC, grant number DEFC52-08NA28616, and by NASA through Einstein Postdoctoral Fellowship grant number PF3-140111 awarded by the Chandra X-ray Center, which is operated by the Smithsonian Astrophysical Observatory for NASA under contract NAS8-03060.

  10. Preparing technicians for engineering materials technology

    NASA Technical Reports Server (NTRS)

    Jacobs, James A.; Metzloff, Carlton H.

    1990-01-01

    A long held principle is that for every engineer and scientist there is a need for ten technicians to maximize the efficiency of the technology team for meeting needs of industry and government. Developing an adequate supply of technicians to meet the requirements of the materials related industry will be a challenge and difficult to accomplish. A variety of agencies feel the need and wish to support development of engineering materials technology programs. In a joint effort among Battelle Laboratories, the Department of Energy (DOE) and Northwest College and University Association for Science (NORCUS), the development of an engineering materials technology program for vocational programs and community colleges for the Pacific Northwest Region was recently completed. This effort has implications for a national model. The model Associate of Applied Science degree in Engineering Materials Technology shown provides a general structure. It purposely has course titles which need delimiting while also including a core of courses necessary to develop cognitive, affective and psychomotor skills with the underlining principles of math, science and technology so students have job entry skills, and so that students can learn about and adapt to evolving technology.

  11. Hospital Guidelines for Diabetes Management and the Joint Commission-American Diabetes Association Inpatient Diabetes Certification.

    PubMed

    Arnold, Pamela; Scheurer, Danielle; Dake, Andrew W; Hedgpeth, Angela; Hutto, Amy; Colquitt, Caroline; Hermayer, Kathie L

    2016-04-01

    The Joint Commission Advanced Inpatient Diabetes Certification Program is founded on the American Diabetes Association's Clinical Practice Recommendations and is linked to the Joint Commission Standards. Diabetes currently affects 29.1 million people in the USA and another 86 million Americans are estimated to have pre-diabetes. On a daily basis at the Medical University of South Carolina (MUSC) Medical Center, there are approximately 130-150 inpatients with a diagnosis of diabetes. The program encompasses all service lines at MUSC. Some important features of the program include: a program champion or champion team, written blood glucose monitoring protocols, staff education in diabetes management, medical record identification of diabetes, a plan coordinating insulin and meal delivery, plans for treatment of hypoglycemia and hyperglycemia, data collection for incidence of hypoglycemia, and patient education on self-management of diabetes. The major clinical components to develop, implement, and evaluate an inpatient diabetes care program are: I. Program management, II. Delivering or facilitating clinical care, III. Supporting self-management, IV. Clinical information management and V. performance measurement. The standards receive guidance from a Disease-Specific Care Certification Advisory Committee, and the Standards and Survey Procedures Committee of the Joint Commission Board of Commissioners. The Joint Commission-ADA Advanced Inpatient Diabetes Certification represents a clinical program of excellence, improved processes of care, means to enhance contract negotiations with providers, ability to create an environment of teamwork, and heightened communication within the organization. Published by Elsevier Inc.

  12. Program Validation: Four Case Studies. A Brief Report on Four Projects and Their Experiences Before the Joint Dissemination Review Panel.

    ERIC Educational Resources Information Center

    Taylor, Nancy, Ed.

    In an effort to clearly illustrate the most effective approach when seeking Joint Dissemination Review Panel (JDRP) validation, this report describes four different educational programs. A program involving prekindergarten special students, and a project entailing a systems approach for disadvantaged elementary school students, were awarded…

  13. Estimating anatomical wrist joint motion with a robotic exoskeleton.

    PubMed

    Rose, Chad G; Kann, Claudia K; Deshpande, Ashish D; O'Malley, Marcia K

    2017-07-01

    Robotic exoskeletons can provide the high intensity, long duration targeted therapeutic interventions required for regaining motor function lost as a result of neurological injury. Quantitative measurements by exoskeletons have been proposed as measures of rehabilitative outcomes. Exoskeletons, in contrast to end effector designs, have the potential to provide a direct mapping between human and robot joints. This mapping rests on the assumption that anatomical axes and robot axes are aligned well, and that movement within the exoskeleton is negligible. These assumptions hold well for simple one degree-of-freedom joints, but may not be valid for multi-articular joints with unique musculoskeletal properties such as the wrist. This paper presents an experiment comparing robot joint kinematic measurements from an exoskeleton to anatomical joint angles measured with a motion capture system. Joint-space position measurements and task-space smoothness metrics were compared between the two measurement modalities. The experimental results quantify the error between joint-level position measurements, and show that exoskeleton kinematic measurements preserve smoothness characteristics found in anatomical measures of wrist movements.

  14. Effects of load proportioning on the capacity of multiple-hole composite joints

    NASA Technical Reports Server (NTRS)

    Hyer, M. W.; Chastain, P. A.

    1985-01-01

    This study addresses the issue of adjusting the proportion of load transmitted by each hole in a multiple-hole joint so that the joint capacity is a maximum. Specifically two-hole-in-series joints are examined. The results indicate that when each hole reacts 50% of the total load, the joint capacity is not a maximum. One hole generally is understressed at joint failure. The algorithm developed to determine the load proportion at each hole which results in maximum capacity is discussed. The algorithm includes two-dimensional finite-element stress analysis and failure criteria. The algorithm is used to study the effects of joint width, hole spacing, and hole to joint-end distance on load proportioning and capacity. To study hole size effects, two hole diameters are considered. Three laminates are considered: a quasi-isotropic laminate; a cross-ply laminate; and a 45 degree angle-ply laminate. By proportioning the load, capacity can be increased generally from 5 to 10%. In some cases a greater increase is possible.

  15. Joint Enrollment: Fiscal Year 2016. Annual Report

    ERIC Educational Resources Information Center

    Division of Community Colleges and Workforce Preparation, Iowa Department of Education, 2017

    2017-01-01

    The Iowa Department of Education collects information on joint enrollment from Iowa's 15 community colleges. Jointly enrolled students are high school students enrolled in community college credit coursework. Most jointly enrolled students enroll through Senior Year Plus programs such as Postsecondary Enrollment Options (PSEO) and concurrent…

  16. Passive Capture Joint with Three Degrees of Freedom

    NASA Technical Reports Server (NTRS)

    Cloyd, Richard A. (Inventor); Weddendorf, Bruce (Inventor)

    2001-01-01

    A passive capture joint with three degrees of freedom is presented wherein two structural elements are joined together solely by moving the two elements into position, and which when joined together have rotation in all three axes. The inventive apparatus is comprised of two halves: (1) a joint ball mounted on a stem as in a common trailer hitch, and; (2) a socket. The socket consists of a base having an exterior wall and forming an interior chamber, the chamber having a top end and a bottom end, and an interior wall. The chamber is open at the top end, and forms a spherical cup at the bottom end. The socket base's interior chamber is sized to accept the joint ball. The base also forms at least one bore at an acute angle away from the interior chamber's open end. The bores have a first opening in the interior wall of the chamber, and a second opening in the exterior wall of the base. Retaining balls sized to fit within the bores, but to only partially pass through the first opening, are moveably housed within the bores. The retaining balls are moveably held in the first opening by a compression spring housed in the bore. As the joint ball is inserted in the chamber it forces the retaining balls back into the bore until the equator of the joint ball passes. Because the bore is at an acute angle to the chamber the joint ball cannot exit the chamber without the joint being unlocked. The joint is unlocked by rotating a locking ring which encircles the base and covers the second opening. The locking ring has a radial slot for each retaining ball, disposed angularly from the base, and sized to allow passage of the retaining ball in the radial direction when the locking ring is rotated to align the radial slot with the second opening.

  17. Systems engineering in a joint program environment: the joint helmet-mounted cueing system

    NASA Astrophysics Data System (ADS)

    Wilkins, Donald F.

    1999-07-01

    The Joint Helmet Mounted Cueing System (JHMCS) is a design program involving two airframe companies (Boeing and Lockheed Martin), two services (USAF and USN) and four aircraft platforms: the F-22, the F-16, the F/A-18 and the F-15. Developing equipment requirements for the combined operational and environmental needs of these diverse communities is a significant challenge. In addition, the team is geographically dispersed which presented challenges in communication and coordination. This paper details the lessons learned in producing a cost-effective design within a short development schedule and makes recommendations for future development programs.

  18. Joint University Program for Air Transportation Research, 1988-1989

    NASA Technical Reports Server (NTRS)

    Morrell, Frederick R. (Compiler)

    1990-01-01

    The research conducted during 1988 to 1989 under the NASA/FAA-sponsored Joint University Program for Air Transportation Research is summarized. The Joint University Program is a coordinated set of three grants sponsored by NASA Langley Research Center and the Federal Aviation Administration, one each with the Massachusetts Institute of Technology, Ohio University, and Princeton University. Completed works, status reports, and annotated bibliographies are presented for research topics, which include computer science, guidance and control theory and practice, aircraft performance, flight dynamics, and applied experimental psychology. An overview of the year's activities for each university is also presented.

  19. The ASTER Global Topographic Data Set

    NASA Astrophysics Data System (ADS)

    Abrams, M.; Bailey, B.; Tsu, H.; Hato, M.

    2009-12-01

    The availability of an up-to-date, high-resolution global digital elevation model (DEM) has been a priority of the Earth observation community for a long time. Until now, the best publicly available global data set has been the 100 m SRTM topography, covering 60 degrees north to 57 degrees south latitude On June 29 Japan’s Ministry of Economy, Trade, and Industry (METI) and the United States National Aeronautics and Space Administration (NASA) released the ASTER Global (GDEM) created from Advanced Spaceborne Thermal Emission and Reflection Radiometer (ASTER) data. ASTER is an imaging instrument built by METI and operating on the NASA Terra platform. ASTER has a backward- looking stereo band, producing stereo pairs in the near-infrared wavelength region; from these stereo pairs, DEMs with 30 m postings (1 arc-second) can be produced. The joint US/Japan ASTER Project completed a program to produce a global DEM (GDEM). The ASTER GDEM was created by stereo-correlating the entire 1,200,000-scene ASTER archive; stacking and averaging the individual DEMs; cloud screening; and filling voids or holes using SRTM 100 m or other data where available. An extensive validation program was completed prior to release of the GDEM. Validation of the GDEM involved comparisons against higher resolution DEMs worldwide by many organizations. Results indicate that globally, the GDEM meets the 20 m vertical accuracy requirement at the 95% confidence level. Accompanying each tile is another data plane indicating the number of individual DEMs that went into the stack, or identifying the data source used to fill the void. At the November 2007 GEO Ministerial Summit, NASA and METI were invited by GEO to contribute this global DEM to GEOSS. Both countries accepted the invitation. Consequently, the ASTER GDEM is offered at no charge to users worldwide. It is packaged in 1 degree-by-1 degree tiles, and covers the Earth’s land surfaces between 83 degree N and 83 degree S latitudes with estimated accuracies of 20 m for vertical data and 30 m for horizontal data. It is distributed by both METI’s Earth Remote Sensing Data Analysis Center organization in Japan, and NASA’s Land Processes Distributed Active Archive Center in the U.S.

  20. Associations of Smoking and Age With Inflammatory Joint Signs Among Unaffected First-Degree Relatives of Rheumatoid Arthritis Patients: Results From Studies of the Etiology of Rheumatoid Arthritis.

    PubMed

    Sparks, Jeffrey A; Chang, Shun-Chiao; Deane, Kevin D; Gan, Ryan W; Kristen Demoruelle, M; Feser, Marie L; Moss, LauraKay; Buckner, Jane H; Keating, Richard M; Costenbader, Karen H; Gregersen, Peter K; Weisman, Michael H; Mikuls, Ted R; O'Dell, James R; Michael Holers, V; Norris, Jill M; Karlson, Elizabeth W

    2016-08-01

    To examine whether genetic, environmental, and serologic rheumatoid arthritis (RA) risk factors are associated with inflammatory joint signs in a cohort of first-degree relatives (FDRs) of RA patients. We evaluated RA risk factors and inflammatory joint signs in a prospective cohort of FDRs without RA in the Studies of the Etiology of RA. Genetic factors included 5 HLA-DRB1 shared epitope alleles and 45 RA-associated single-nucleotide polymorphisms; loci were combined using genetic risk scores weighted by RA risk. Environmental factors (smoking, body mass index, education, and parity) and RA-related autoantibodies were assessed at baseline. Physical examination was performed at baseline and 2-year follow-up, by observers who were blinded with regard to autoantibody status, to assess inflammatory joint signs as tender or swollen joints at sites typical for RA. Logistic regression was performed to evaluate associations of genetic, environmental, and serologic factors with inflammatory joint signs. We analyzed 966 non-Hispanic white FDRs at baseline and 262 at 2-year follow-up after excluding those with inflammatory joint signs at baseline. The mean ± SD age was 47.2 ± 15.5 years, 71% were female, and 55% were shared epitope positive. Smoking >10 pack-years was associated with inflammatory joint signs at baseline (odds ratio [OR] 1.89 [95% confidence interval (95% CI) 1.26-2.82]) and at 2 years (OR 2.66 [95% CI 1.01-7.03]), compared to never smokers. There was a significant interaction between smoking and age with regard to risk of inflammatory joint signs (P = 0.02). FDRs younger than 50 years with >10 pack-years had the highest risk of inflammatory joint signs (OR 4.39 [95% CI 2.22-8.66], compared to never smokers younger than 50 years). In a high-risk cohort of FDRs, smoking and age were associated with both prevalent and incident inflammatory joint signs at sites typical for RA. Further prospective investigations of the factors affecting the transitions between preclinical RA phases are warranted. © 2016, American College of Rheumatology.

  1. Stochastic estimation of human shoulder impedance with robots: an experimental design.

    PubMed

    Park, Kyungbin; Chang, Pyung Hun

    2011-01-01

    Previous studies assumed the shoulder as a hinge joint during human arm impedance measurement. This is obviously a vast simplification since the shoulder is a complex of several joints with multiple degrees of freedom. In the present work, a practical methodology for more general and realistic estimation of human shoulder impedance is proposed and validated with a spring array. It includes a gravity compensation scheme, which is developed and used for the experiments with a spatial three degrees of freedom PUMA-type robot. The experimental results were accurate and reliable, and thus it has shown a strong potential of the proposed methodology in the estimation of human shoulder impedance. © 2011 IEEE

  2. Autonomous manipulation on a robot: Summary of manipulator software functions

    NASA Technical Reports Server (NTRS)

    Lewis, R. A.

    1974-01-01

    A six degree-of-freedom computer-controlled manipulator is examined, and the relationships between the arm's joint variables and 3-space are derived. Arm trajectories using sequences of third-degree polynomials to describe the time history of each joint variable are presented and two approaches to the avoidance of obstacles are given. The equations of motion for the arm are derived and then decomposed into time-dependent factors and time-independent coefficients. Several new and simplifying relationships among the coefficients are proven. Two sample trajectories are analyzed in detail for purposes of determining the most important contributions to total force in order that relatively simple approximations to the equations of motion can be used.

  3. 77 FR 70914 - Federal Motor Vehicle Safety Standards; Side Impact Protection

    Federal Register 2010, 2011, 2012, 2013, 2014

    2012-11-28

    ..., S12.3.2(c), S12.3.3(c) and S12.3.4(l), the `` '' symbols for the discrete arm position settings were... centerline. The shoulder-arm joint allows for discrete arm positions at 0, 45, 90, 135, and 180 degree... thoracic spine centerline. The shoulder-arm joint allows for discrete arm positions at 0, [[Page 70915

  4. Type synthesis and preliminary design of devices supporting lower limb's rehabilitation.

    PubMed

    Olinski, Michał; Lewandowski, Bogusz; Gronowicz, Antoni

    2015-01-01

    Based on the analysis of existing solutions, biomechanics of human lower limbs and anticipated applications, results of con- siderations concerning the necessary number of degrees of freedom for the designed device supporting rehabilitation of lower extremities are presented. An analysis was carried out in order to determine the innovative kinematic structure of the device, ensuring sufficient mobility and functionality while minimizing the number of degrees of freedom. With the aid of appropriate formalised meth- ods, for instance, type synthesis, a complete variety of solutions for leg joints were obtained in the form of basic and kinematic schemes, having the potential to find application in devices supporting lower limb rehabilitation. A 3D model of ankle joint module was built in Autodesk Inventor System, then imported to Adams and assembled into a moving numerical model of a mechanism. Several conducted simulations resulted in finding the required maximum stroke of the cylinders. A comparison of the angular ranges of ankle joint and similar devices with the ones achieved by the designed device indicated a sufficient reserve allowing not only movements typical of gait, but approximately achieving the passive range of motion for the ankle joint.

  5. The mechanism of joint capsule thermal modification in an in-vitro sheep model.

    PubMed

    Hayashi, K; Peters, D M; Thabit, G; Hecht, P; Vanderby, R; Fanton, G S; Markel, M D

    2000-01-01

    The purpose of this study was to understand the mechanism responsible for joint capsule shrinkage after nonablative laser application in an in-vitro sheep model. Femoropatellar joint capsular tissue specimens harvested from 20 adult sheep were treated with one of three power settings of a holmium:yttrium-aluminum-garnet laser or served as a control. Laser treatment significantly shortened the tissue and decreased tissue stiffness in all three laser groups, whereas failure strength was not altered significantly by laser treatment. Transmission electron microscopic examination showed swollen collagen fibrils and loss of membrane integrity of fibroblasts. A thermometric study revealed nonablative laser energy caused tissue temperature to rise in the range of 64 degrees C to 100 degrees C. Electrophoresis after trypsin digestion of the tissue revealed significant loss of distinct alpha bands of Type I collagen in laser treated samples, whereas alpha bands were present in laser treated tissue without trypsin digestion. The results of this study support the concept that the primary mechanism responsible for the effect of nonablative laser energy is thermal denaturation of collagen in joint capsular tissue associated with unwinding of the triple helical structure of the collagen molecule.

  6. Effects of a 6-week, individualized, supervised exercise program for people with bleeding disorders and hemophilic arthritis.

    PubMed

    Mulvany, Ruth; Zucker-Levin, Audrey R; Jeng, Michael; Joyce, Catherine; Tuller, Janet; Rose, Jonathan M; Dugdale, Marion

    2010-04-01

    People with bleeding disorders may develop severe arthritis due to joint hemorrhages. Exercise is recommended for people with bleeding disorders, but guidelines are vague and few studies document efficacy. In this study, 65% of people with bleeding disorders surveyed reported participating in minimal exercise, and 50% indicated a fear of exercise-induced bleeding, pain, or physical impairment. The purpose of this study was to examine the feasibility, safety, and efficacy of a professionally designed, individualized, supervised exercise program for people with bleeding disorders. A single-group, pretest-posttest clinical design was used. Thirty-three patients (3 female, 30 male; 7-57 years of age) with mild to severe bleeding disorders were enrolled in the study. Twelve patients had co-existing illnesses, including HIV/AIDS, hepatitis, diabetes, fibromyalgia, neurofibromatosis, osteopenia, osteogenesis imperfecta, or cancer. Pre- and post-program measures included upper- and lower-extremity strength (force-generating capacity), joint range of motion, joint and extremity circumference, and distance walked in 6 minutes. Each patient was prescribed a 6-week, twice-weekly, individualized, supervised exercise program. Twenty participants (61%) completed the program. Pre- and post-program data were analyzed by paired t tests for all participants who completed the program. No exercise-induced injuries, pain, edema, or bleeding episodes were reported. Significant improvements occurred in joint motion, strength, and distance walked in 6 minutes, with no change in joint circumference. The greatest gains were among the individuals with the most severe joint damage and coexisting illness. Limitations included a small sample size with concomitant disease, which is common to the population, and a nonblinded examiner. A professionally designed and supervised, individualized exercise program is feasible, safe, and beneficial for people with bleeding disorders, even in the presence of concomitant disease. A longitudinal study with a larger sample size, a blinded examiner, and a control group is needed to confirm the results.

  7. 75 FR 32535 - ITS Joint Program Office; IntelliDriveSM

    Federal Register 2010, 2011, 2012, 2013, 2014

    2010-06-08

    ... DEPARTMENT OF TRANSPORTATION ITS Joint Program Office; IntelliDrive\\SM\\ Deployment Scenarios Workshop; Notice of Workshop AGENCY: Research and Innovative Technology Administration, Department of... participants to identify advantages and disadvantages of each of the draft scenarios and critical policy and...

  8. Are you ready? Managing transportation resources through the Y2K weekend

    DOT National Transportation Integrated Search

    2011-01-01

    The Clarus Initiative, a joint effort of the U.S. Department of Transportation Intelligent Transportation Systems (ITS) Joint Program Office and FHWAs Road Weather Management Program (RWMP), is a six-year effort to develop and demonstrate an integ...

  9. Design of Revolute Joints for In-Mold Assembly Using Insert Molding.

    PubMed

    Ananthanarayanan, Arvind; Ehrlich, Leicester; Desai, Jaydev P; Gupta, Satyandra K

    2011-12-01

    Creating highly articulated miniature structures requires assembling a large number of small parts. This is a very challenging task and increases cost of mechanical assemblies. Insert molding presents the possibility of creating a highly articulated structure in a single molding step. This can be accomplished by placing multiple metallic bearings in the mold and injecting plastic on top of them. In theory, this idea can generate a multi degree of freedom structures in just one processing step without requiring any post molding assembly operations. However, the polymer material has a tendency to shrink on top of the metal bearings and hence jam the joints. Hence, until now insert molding has not been used to create articulated structures. This paper presents a theoretical model for estimating the extent of joint jamming that occurs due to the shrinkage of the polymer on top of the metal bearings. The level of joint jamming is seen as the effective torque needed to overcome the friction in the revolute joints formed by insert molding. We then use this model to select the optimum design parameters which can be used to fabricate functional, highly articulating assemblies while meeting manufacturing constraints. Our analysis shows that the strength of weld-lines formed during the in-mold assembly process play a significant role in determining the minimum joint dimensions necessary for fabricating functional revolute joints. We have used the models and methods described in this paper to successfully fabricate the structure for a minimally invasive medical robot prototype with potential applications in neurosurgery. To the best of our knowledge, this is the first demonstration of building an articulated structure with multiple degrees of freedom using insert molding.

  10. Differences in the stress distribution in the distal femur between patellofemoral joint replacement and total knee replacement: a finite element study

    PubMed Central

    2012-01-01

    Background Patellofemoral joint replacement is a successful treatment option for isolated patellofemoral osteoarthritis. However, results of later conversion to total knee replacement may be compromised by periprosthetic bone loss. Previous clinical studies have demonstrated a decrease in distal femoral bone mineral density after patellofemoral joint replacement. It is unclear whether this is due to periprosthetic stress shielding. The main objective of the current study was to evaluate the stress shielding effect of prosthetic replacement with 2 different patellofemoral prosthetic designs and with a total knee prosthesis. Methods We developed a finite element model of an intact patellofemoral joint, and finite element models of patellofemoral joint replacement with a Journey PFJ prosthesis, a Richards II prosthesis, and a Genesis II total knee prosthesis. For each of these 4 finite element models, the average Von Mises stress in 2 clinically relevant regions of interest were evaluated during a simulated squatting movement until 120 degrees of flexion. Results During deep knee flexion, in the anterior region of interest, the average Von Mises stress with the Journey PFJ design was comparable to the physiological knee, while reduced by almost 25% for both the Richards II design and the Genesis II total knee joint replacement design. The average Von Mises stress in the supracondylar region of interest was similar for both patellofemoral prosthetic designs and the physiological model, with slightly lower stress for the Genesis II design. Conclusions Patellofemoral joint replacement results in periprosthetic stress-shielding, although to a smaller degree than in total knee replacement. Specific patellofemoral prosthetic design properties may result in differences in femoral stress shielding. PMID:22704638

  11. Cluster-based upper body marker models for three-dimensional kinematic analysis: Comparison with an anatomical model and reliability analysis.

    PubMed

    Boser, Quinn A; Valevicius, Aïda M; Lavoie, Ewen B; Chapman, Craig S; Pilarski, Patrick M; Hebert, Jacqueline S; Vette, Albert H

    2018-04-27

    Quantifying angular joint kinematics of the upper body is a useful method for assessing upper limb function. Joint angles are commonly obtained via motion capture, tracking markers placed on anatomical landmarks. This method is associated with limitations including administrative burden, soft tissue artifacts, and intra- and inter-tester variability. An alternative method involves the tracking of rigid marker clusters affixed to body segments, calibrated relative to anatomical landmarks or known joint angles. The accuracy and reliability of applying this cluster method to the upper body has, however, not been comprehensively explored. Our objective was to compare three different upper body cluster models with an anatomical model, with respect to joint angles and reliability. Non-disabled participants performed two standardized functional upper limb tasks with anatomical and cluster markers applied concurrently. Joint angle curves obtained via the marker clusters with three different calibration methods were compared to those from an anatomical model, and between-session reliability was assessed for all models. The cluster models produced joint angle curves which were comparable to and highly correlated with those from the anatomical model, but exhibited notable offsets and differences in sensitivity for some degrees of freedom. Between-session reliability was comparable between all models, and good for most degrees of freedom. Overall, the cluster models produced reliable joint angles that, however, cannot be used interchangeably with anatomical model outputs to calculate kinematic metrics. Cluster models appear to be an adequate, and possibly advantageous alternative to anatomical models when the objective is to assess trends in movement behavior. Copyright © 2018 Elsevier Ltd. All rights reserved.

  12. Does distal tibiofibular joint mobilization decrease limitation of ankle dorsiflexion?

    PubMed

    Fujii, Misaki; Suzuki, Daisuke; Uchiyama, Eiichi; Muraki, Takayuki; Teramoto, Atsushi; Aoki, Mitsuhiro; Miyamoto, Shigenori

    2010-02-01

    Limitation of ankle motion is in many cases treated by joint mobilization (JM), a kind of manual physical therapy technique. Until now, the JM approach has mainly focused on the talocrural joint, with less attention to the distal tibiofibular joint. We applied cyclic loading to the lateral malleolus as in JM in order to clarify the relationship between the dorsiflexion angle and the excursion of the lateral malleolus. Seven normal, fresh-frozen cadaver legs were used. To each specimen, cyclic loading with a 30N force was applied 1000 times to the lateral malleolus at a speed of 15N/s. The displacement of the lateral malleolus was measured with a magnetic tracking system. The maximum dorsiflexion angle was measured before and after cyclic loading. After the first 100 and 1000 times of cyclic loading, the tibia was displaced 0.44+/-0.30mm and 0.75+/-0.36mm, respectively, and the fibula was displaced 0.44+/-0.28mm and 0.92+/-0.39mm, respectively. The average dorsiflexion angle increased from 14.36+/-7.51 degrees to 16.74+/-7.21 degrees after cyclic loading (P<0.05). Movement of the distal tibiofibular joint led to a significant increase in the range of ankle dorsiflexion. These results suggest that tibiofibular JM would be effective for limitation of ankle dorsiflexion.

  13. Japan-U.S. Joint Ventures in Higher Education: Language Education in an Uncertain Future.

    ERIC Educational Resources Information Center

    Clayton, Thomas

    1991-01-01

    Examines Japanese-U.S. joint venture language or U.S.-style education programs in Japan. These programs offer language and cultural education classes for those interested in English and for students who have failed in the Japanese education system. Problems facing these programs and the need to explore new, English-language education markets are…

  14. Torque Control of Underactuated Tendon-driven Robotic Fingers

    NASA Technical Reports Server (NTRS)

    Ihrke, Chris A. (Inventor); Wampler, Charles W. (Inventor); Abdallah, Muhammad E. (Inventor); Reiland, Matthew J. (Inventor); Diftler, Myron A. (Inventor); Bridgwater, Lyndon (Inventor); Platt, Robert (Inventor)

    2013-01-01

    A robotic system includes a robot having a total number of degrees of freedom (DOF) equal to at least n, an underactuated tendon-driven finger driven by n tendons and n DOF, the finger having at least two joints, being characterized by an asymmetrical joint radius in one embodiment. A controller is in communication with the robot, and controls actuation of the tendon-driven finger using force control. Operating the finger with force control on the tendons, rather than position control, eliminates the unconstrained slack-space that would have otherwise existed. The controller may utilize the asymmetrical joint radii to independently command joint torques. A method of controlling the finger includes commanding either independent or parameterized joint torques to the controller to actuate the fingers via force control on the tendons.

  15. 40 CFR 80.1161 - Who is liable for violations under the RFS program?

    Code of Federal Regulations, 2010 CFR

    2010-07-01

    ... subpart that is committed by any of its subsidiaries. (d) Joint venture liability. Each partner to a joint venture is jointly and severally liable for any violation of this subpart that is committed by the joint venture operation. [72 FR 24003, May 1, 2007] ...

  16. Corona performance of a compact 230-kV line

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chartier, V.L.; Blair, D.E.; Easley, M.D.

    Permitting requirements and the acquisition of new rights-of-way for transmission facilities has in recent years become increasingly difficult for most utilities, including Puget Sound Power and Light Company. In order to maintain a high degree of reliability of service while being responsive to public concerns regarding the siting of high voltage (HV) transmission facilities, Puget Power has found it necessary to more heavily rely upon the use of compact lines in franchise corridors. Compaction does, however, precipitate increased levels of audible noise (AN) and radio and TV interference (RI and TVI) due to corona on the conductors and insulator assemblies.more » Puget Power relies upon the Bonneville Power Administration (BPA) Corona and Field Effects computer program to calculate AN and RI for new lines. Since there was some question of the program`s ability to accurately represent quiet 230-kV compact designs, a joint project was undertaken with BPA to verify the program`s algorithms. Long-term measurements made on an operating Puget Power 230-kV compact line confirmed the accuracy of BPA`s AN model; however, the RI measurements were much lower than predicted by the BPA and other programs. This paper also describes how the BPA computer program can be used to calculate the voltage needed to expose insulator assemblies to the correct electric field in single test setups in HV laboratories.« less

  17. Corona performance of a compact 230-kV line

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chartier, V.L.; Blair, D.E.; Easley, M.D.

    Permitting requirements and the acquisition of new rights-of-way for transmission facilities has in recent years become increasingly difficult for most utilities, including Puget Sound Power and Light Company. In order to maintain a high degree of reliability of service while being responsive to public concerns regarding the siting of high voltage (HV) transmission facilities, Puget Power has found it necessary to more heavily rely upon the use of compact lines in franchise corridors. Compaction does, however, precipitant increased levels of audible noise (AN) and radio and TV interference (RI and TVI) due to corona on the conductors and insulator assemblies.more » Puget Power relies upon the Bonneville Power Administration (BPA) Corona and Field Effects computer program to calculate AN and RI for new lines. Since there was some question of the program`s ability to accurately represent quiet 230-kV compact designs, a joint project was undertaken with BPA to verify the program`s algorithms. Long-term measurements made on an operating Puget Power 230-kV compact line confirmed the accuracy of BPA`s AN model; however, the RI measurements were much lower than predicted by the BPA computer and other programs. This paper also describes how the BPA computer program can be used to calculate the voltage needed to expose insulator assemblies to the correct electric field in single test setups in HV laboratories.« less

  18. A Systematic Review of Different Implants and Approaches for Proximal Interphalangeal Joint Arthroplasty.

    PubMed

    Yamamoto, Michiro; Malay, Sunitha; Fujihara, Yuki; Zhong, Lin; Chung, Kevin C

    2017-05-01

    Outcomes after implant arthroplasty for primary degenerative and posttraumatic osteoarthritis of the proximal interphalangeal joint were different according to the implant design and surgical approach. The purpose of this systematic review was to evaluate outcomes of various types of implant arthroplasty for proximal interphalangeal joint osteoarthritis, with an emphasis on different surgical approaches. The authors searched all available literature in the PubMed and EMBASE databases for articles reporting on outcomes of implant arthroplasty for proximal interphalangeal joint osteoarthritis. Data collection included active arc of motion, extension lag, and complications. The authors combined the data of various types of surface replacement arthroplasty into one group for comparison with silicone arthroplasty. A total of 849 articles were screened, yielding 40 studies for final review. The mean postoperative arc of motion and the mean gain in arc of motion of silicone implant with the volar approach were 58 and 17 degrees, respectively, which was greater than surface replacement implant with the dorsal approach at 51 and 8 degrees, respectively. The mean postoperative extension lag of silicone implant with the volar approach and surface replacement with the dorsal approach was 5 and 14 degrees, respectively. The revision rate of silicone implant with the volar approach and surface replacement with the dorsal approach was 6 percent and 18 percent at a mean follow-up of 41.2 and 51 months, respectively. Silicone implant with the volar approach showed the best arc of motion, with less extension lag and fewer complications after surgery among all the implant designs and surgical approaches.

  19. Robotic Arm Manipulator Using Active Control for Sample Acquisition and Transfer, and Passive Mode for Surface Compliance

    NASA Technical Reports Server (NTRS)

    Liu, Jun; Underhill, Michael L.; Trease, Brian P.; Lindemann, Randel A.

    2010-01-01

    A robotic arm that consists of three joints with four degrees of freedom (DOF) has been developed. It can carry an end-effector to acquire and transfer samples by using active control and comply with surface topology in a passive mode during a brief surface contact. The three joints are arranged in such a way that one joint of two DOFs is located at the shoulder, one joint of one DOF is located at the elbow, and one joint of one DOF is located at the wrist. Operationally, three DOFs are moved in the same plane, and the remaining one on the shoulder is moved perpendicular to the other three for better compliance with ground surface and more flexibility of sample handling. Three out of four joints are backdriveable, making the mechanism less complex and more cost effective

  20. Control of joint motion simulators for biomechanical research

    NASA Technical Reports Server (NTRS)

    Colbaugh, R.; Glass, K.

    1992-01-01

    The authors present a hierarchical adaptive algorithm for controlling upper extremity human joint motion simulators. A joint motion simulator is a computer-controlled, electromechanical system which permits the application of forces to the tendons of a human cadaver specimen in such a way that the cadaver joint under study achieves a desired motion in a physiologic manner. The proposed control scheme does not require knowledge of the cadaver specimen dynamic model, and solves on-line the indeterminate problem which arises because human joints typically possess more actuators than degrees of freedom. Computer simulation results are given for an elbow/forearm system and wrist/hand system under hierarchical control. The results demonstrate that any desired normal joint motion can be accurately tracked with the proposed algorithm. These simulation results indicate that the controller resolved the indeterminate problem redundancy in a physiologic manner, and show that the control scheme was robust to parameter uncertainty and to sensor noise.

  1. Long-Term Pavement Performance Automated Faulting Measurement

    DOT National Transportation Integrated Search

    2015-02-01

    This study focused on identifying transverse joint locations on jointed plain concrete pavements using an automated joint detection algorithm and computing faulting at these locations using Long-Term Pavement Performance (LTPP) Program profile data c...

  2. Can generic knee joint models improve the measurement of osteoarthritic knee kinematics during squatting activity?

    PubMed

    Clément, Julien; Dumas, Raphaël; Hagemeister, Nicola; de Guise, Jaques A

    2017-01-01

    Knee joint kinematics derived from multi-body optimisation (MBO) still requires evaluation. The objective of this study was to corroborate model-derived kinematics of osteoarthritic knees obtained using four generic knee joint models used in musculoskeletal modelling - spherical, hinge, degree-of-freedom coupling curves and parallel mechanism - against reference knee kinematics measured by stereo-radiography. Root mean square errors ranged from 0.7° to 23.4° for knee rotations and from 0.6 to 9.0 mm for knee displacements. Model-derived knee kinematics computed from generic knee joint models was inaccurate. Future developments and experiments should improve the reliability of osteoarthritic knee models in MBO and musculoskeletal modelling.

  3. Scope of Work for Joint W28/AGM-28A Missile Firing Program

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Bice, R.A.

    1964-12-04

    List of principles & responsibilities of this Sandia/AEC & Air Force joint program for verifying the functionability of the W28/AGM-28A/B-52 weapon system under actual routine handling, assembling, and firing during a stockpile-to-target sequence.

  4. Scope of Work for the Joint W53/Titan II Missile Firing Program

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Not Available

    1964-10-28

    List of principles & responsibilities of this Sandia/AEC & Air Force joint program for verifying the functionability of the W28/AGM-28A/B-52 weapon system under actual routine handling, assembling, and firing during a stockpile-to-target sequence.

  5. BUILDING MODEL ANALYSIS APPLICATIONS WITH THE JOINT UNIVERSAL PARAMETER IDENTIFICATION AND EVALUATION OF RELIABILITY (JUPITER) API

    EPA Science Inventory

    The open-source, public domain JUPITER (Joint Universal Parameter IdenTification and Evaluation of Reliability) API (Application Programming Interface) provides conventions and Fortran-90 modules to develop applications (computer programs) for analyzing process models. The input ...

  6. Integrative Curriculum Development in Nuclear Education and Research Vertical Enhancement Program

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Egarievwe, Stephen U.; Jow, Julius O.; Edwards, Matthew E.

    Using a vertical education enhancement model, a Nuclear Education and Research Vertical Enhancement (NERVE) program was developed. The NERVE program is aimed at developing nuclear engineering education and research to 1) enhance skilled workforce development in disciplines relevant to nuclear power, national security and medical physics, and 2) increase the number of students and faculty from underrepresented groups (women and minorities) in fields related to the nuclear industry. The program uses multi-track training activities that vertically cut across the several education domains: undergraduate degree programs, graduate schools, and post-doctoral training. In this paper, we present the results of an integrativemore » curriculum development in the NERVE program. The curriculum development began with nuclear content infusion into existing science, engineering and technology courses. The second step involved the development of nuclear engineering courses: 1) Introduction to Nuclear Engineering, 2) Nuclear Engineering I, and 2) Nuclear Engineering II. The third step is the establishment of nuclear engineering concentrations in two engineering degree programs: 1) electrical engineering, and 2) mechanical engineering. A major outcome of the NERVE program is a collaborative infrastructure that uses laboratory work, internships at nuclear facilities, on-campus research, and mentoring in collaboration with industry and government partners to provide hands-on training for students. The major activities of the research and education collaborations include: - One-week spring training workshop at Brookhaven National Laboratory: The one-week training and workshop is used to enhance research collaborations and train faculty and students on user facilities/equipment at Brookhaven National Laboratory, and for summer research internships. Participants included students, faculty members at Alabama A and M University and research collaborators at BNL. The activities include 1) tour and introduction to user facilities/equipment at BNL that are used for research in room-temperature semiconductor nuclear detectors, 2) presentations on advances on this project and on wide band-gap semiconductor nuclear detectors in general, and 3) graduate students' research presentations. - Invited speakers and lectures: This brings collaborating research scientist from BNL to give talks and lectures on topics directly related to the project. Attendance includes faculty members, researchers and students throughout the university. - Faculty-students team summer research at BNL: This DOE and National Science Foundation (NSF) program help train students and faculty members in research. Faculty members go on to establish research collaborations with scientists at BNL, develop and submit research proposals to funding agencies, transform research experience at BNL to establish and enhance reach capabilities at home institution, and integrate their research into teaching through class projects and hands-on training for students. The students go on to participate in research work at BNL and at home institution, co-author research papers for conferences and technical journals, and transform their experiences into developing senior and capstone projects. - Grant proposal development: Faculty members in the NERVE program collaborate with BNL scientists to develop proposals, which often help to get external funding needed to expand and sustain the continuity of research activities and supports for student's wages and scholarships (stipends, tuition and fees). - Faculty development and mentoring: The above collaboration activities help faculty professional development. The experiences, grants, joint publications in technical journals, and supervision of student's research, including thesis and dissertation research projects, contribute greatly to faculty development. Senior scientists at BNL and senior faculty members on campus jointly mentor junior faculty members to enhance their professional growth. - Graduate thesis and dissertation research: Brookhaven National Laboratory provides unique opportunities and outstanding research resources for the NERVE program graduate research. Scientists from BNL serve in master's degree thesis and PhD dissertation committees, where they play active roles in the supervision of the research. (authors)« less

  7. In-vivo measurement of dynamic joint motion using high speed biplane radiography and CT: application to canine ACL deficiency.

    PubMed

    Tashman, Scott; Anderst, William

    2003-04-01

    Dynamic assessment of three-dimensional (3D) skeletal kinematics is essential for understanding normal joint function as well as the effects of injury or disease. This paper presents a novel technique for measuring in-vivo skeletal kinematics that combines data collected from high-speed biplane radiography and static computed tomography (CT). The goals of the present study were to demonstrate that highly precise measurements can be obtained during dynamic movement studies employing high frame-rate biplane video-radiography, to develop a method for expressing joint kinematics in an anatomically relevant coordinate system and to demonstrate the application of this technique by calculating canine tibio-femoral kinematics during dynamic motion. The method consists of four components: the generation and acquisition of high frame rate biplane radiographs, identification and 3D tracking of implanted bone markers, CT-based coordinate system determination, and kinematic analysis routines for determining joint motion in anatomically based coordinates. Results from dynamic tracking of markers inserted in a phantom object showed the system bias was insignificant (-0.02 mm). The average precision in tracking implanted markers in-vivo was 0.064 mm for the distance between markers and 0.31 degree for the angles between markers. Across-trial standard deviations for tibio-femoral translations were similar for all three motion directions, averaging 0.14 mm (range 0.08 to 0.20 mm). Variability in tibio-femoral rotations was more dependent on rotation axis, with across-trial standard deviations averaging 1.71 degrees for flexion/extension, 0.90 degree for internal/external rotation, and 0.40 degree for varus/valgus rotation. Advantages of this technique over traditional motion analysis methods include the elimination of skin motion artifacts, improved tracking precision and the ability to present results in a consistent anatomical reference frame.

  8. Identifying the control structure of multijoint coordination during pistol shooting.

    PubMed

    Scholz, J P; Schöner, G; Latash, M L

    2000-12-01

    The question of degrees of freedom in the control of multijoint movement is posed as the problem of discovering how the motor control system constrains the many possible combinations of joint postures to stabilize task-dependent essential variables. Success at a task can be achieved, in principle, by always adopting a particular joint combination. In contrast, we propose a more selective control strategy: variations of the joint configuration that leave the values of essential task variables unchanged are predicted to be less controlled (i.e., stabilized to a lesser degree) than joint configuration changes that shift the values of the task variables. Our experimental task involved shooting with a laser pistol at a target under four conditions. The seven joint angles of the arm were obtained from the recorded positions of markers on the limb segments. The joint configurations observed at each point in normalized time were analyzed with respect to trial-to-trial variability. Different hypotheses about relevant task variables were used to define sets of joint configurations ("uncontrolled manifolds" or UCMs) that, if realized, would leave essential task variables unchanged. The variability of joint configurations was decomposed into components lying parallel to those sets and components lying in their complement. The orientation of the gun's barrel relative to a vector pointing from the gun to the target was the task variable most successful at showing a difference between the two components of joint variability. This variable determines success at the task. Throughout the movement, not only while the gun was pointing at the target, fluctuations of joint configuration that affected this variable were much reduced compared with fluctuations that did not affect this variable. The UCM principle applied to relative gun orientation thus captures the structure of the motor control system across different parts of joint configuration space as the movement evolves in time. This suggests a specific control strategy in which changes of joint configuration that are irrelevant to success at the task are selectively released from control. By contrast, constraints representing an invariant spatial position of the gun or of the arm's center of mass structured joint configuration variability in the early and mid-portion of the movement trajectory, but not at the time of shooting. This specific control strategy is not trivial, because a target can be hit successfully also by controlling irrelevant directions in joint space equally to relevant ones. The results indicate that the method can be successfully used to determine the structure of coordination in joint space that underlies the control of the essential variables for a given task.

  9. Joint University Program for Air Transportation Research, 1983

    NASA Technical Reports Server (NTRS)

    Morrell, Frederick R. (Compiler)

    1987-01-01

    The research conducted during 1983 under the NASA/FAA sponsored Joint University Program for Air Transportation Research is summarized. The material was presented at a conference held at the Federal Aviation Administration Technical Center, Altantic City, New Jersey, December 16, 1983. The Joint University Program is a coordinated set of three grants sponsored by NASA Langley Research Center and the Federal Aviation Administration, one each with the Massachusetts Institute of Technology, Ohio University, and Princeton University. Completed works, status reports, and bibliographies are presented for research topics, which include navigation, guidance, control, and display concepts. An overview of the year's activities for each of the universities is also presented.

  10. Influence of bi- and tri-compartmental knee arthroplasty on the kinematics of the knee joint

    PubMed Central

    2011-01-01

    Background The cruciate ligaments are important stabilizers of the knee joint and determine joint kinematics in the natural knee and after cruciate retaining arthroplasty. No in vitro data is available to biomechanically evaluate the ability of the anterior cruciate ligament (ACL) to maintain knee joint kinematics after bicruciate-retaining bi-compartmental knee arthroplasty (BKA). Therefore, the objective of the current study was to investigate the kinematics of the natural knee joint, before and after installing bicruciate-retaining BKA and posterior cruciate retaining total knee arthroplasty. Specifically, we incorporated a dynamic knee simulator to simulate weight-bearing flexions on cadaveric knee specimen before and after surgical manipulations. Methods In this cadaveric study we investigated rotational and translational tibiofemoral kinematics during simulated weight-bearing flexions of the intact knee, after bi-compartmental knee arthroplasty (BKA+), after resecting the ACL in BKA (BKA-), and after posterior cruciate retaining total knee arthroplasty (TKA). Results Rotation of BKA+ is closest to the intact knee joint, whereas TKA shows significant differences from 30 to 90 degree of flexion. Within the tested flexion range (15 to 90 degree of flexion), there was no significant difference in the anterior-posterior translation among intact, BKA+, and TKA knees. Resecting the ACL in BKA leads to a significant anterior tibial translation. Conclusions BKA with intact cruciate ligaments resembles rotation and translation of the natural knee during a simulated weight-bearing flexion. It is a suitable treatment option for medial and patellofemoral osteoarthritis with advantages in rotational characteristics compared to TKA. PMID:21272328

  11. "APEC blue"--The effects and implications of joint pollution prevention and control program.

    PubMed

    Wang, Hongbo; Zhao, Laijun; Xie, Yujing; Hu, Qingmi

    2016-05-15

    To ensure good air quality in Beijing during Asia-Pacific Economic Cooperation (APEC) China 2014, Beijing and its neighboring five provinces and the associated cities were combined under the Joint Prevention and Control of Atmospheric Pollution (JPCAP) program, which implemented rigorous cooperative emission reduction measures. The program was a unique and large-scale artificial experiment that showed that such measures can achieve excellent results, and it led to the popular "APEC blue" catchphrase (i.e., Beijing's skies became blue as pollution levels decreased). This artificial experiment provided the means to effectively conduct JPCAP strategies in the future. Accordingly, our research focused on the characteristics of the six primary pollutants in Beijing. We found that the JPCAP measures directly reduced concentrations of all pollutants except O3. Through correlation analysis, we found that the band distribution of the cities with strong correlations in PM2.5 and PM10 concentrations was affected by wind conditions. Therefore, JPCAP measures should account for specific seasonal and climatic conditions. Based on cluster analysis using the results from the correlation analysis, we divided 13 cities within a 300-km radius of Beijing into different groups according to the similarity of their PM2.5 and PM10 correlation coefficients. For JPCAP measures relevant to PM2.5 and PM10, we found differences in the degrees of collaboration among cities. Therefore, depending upon the pollutant type, the JPCAP strategy should account for the cities involved, the scope of the core area, and the optimal cities to involve in the collaborative efforts based on cost-effectiveness and collaborative difficulty among the involved cities. Copyright © 2016 Elsevier B.V. All rights reserved.

  12. Metal-ceramic joint assembly

    DOEpatents

    Li, Jian

    2002-01-01

    A metal-ceramic joint assembly in which a brazing alloy is situated between metallic and ceramic members. The metallic member is either an aluminum-containing stainless steel, a high chromium-content ferritic stainless steel or an iron nickel alloy with a corrosion protection coating. The brazing alloy, in turn, is either an Au-based or Ni-based alloy with a brazing temperature in the range of 9500 to 1200.degree. C.

  13. The importance of position and path repeatability on force at the knee during six-DOF joint motion.

    PubMed

    Darcy, Shon P; Gil, Jorge E; Woo, Savio L-Y; Debski, Richard E

    2009-06-01

    Mechanical devices, such as robotic manipulators have been designed to measure joint and ligament function because of their ability to position a diarthrodial joint in six degrees-of-freedom with fidelity. However, the precision and performance of these testing devices vary. Therefore, the objective of this study was to determine the effect of systematic errors in position and path repeatability of two high-payload robotic manipulators (Manipulators 1 and 2) on the resultant forces at the knee. Using a porcine knee, the position and path repeatability of these manipulators were determined during passive flexion-extension with a coordinate measuring machine. The position repeatability of Manipulator 1 was 0.3 mm in position and 0.2 degrees in orientation while Manipulator 2 had a better position repeatability of 0.1 mm in position and 0.1 degrees in orientation throughout the range of positions examined. The corresponding variability in the resultant force at the knee for these assigned positions was 32+/-33 N for Manipulator 1 and 4+/-1 N for Manipulator 2. Furthermore, the repeatability of the trajectory of each manipulator while moving between assigned positions (path repeatability) was 0.8 mm for Manipulator 1 while the path repeatability for Manipulator 2 was improved (0.1 mm). These path discrepancies produced variability in the resultant force at the knee of 44+/-24 and 21+/-8 N, respectively, for Manipulators 1 and 2 primarily due to contact between the articular surfaces of the tibia and femur. Therefore, improved position and path repeatability yields lower variability in the resultant forces at the knee. Although position repeatability has been the most common criteria for evaluating biomechanical testing devices, the current study has clearly demonstrated that path repeatability can have an even larger effect on the variability in resultant force at the knee. Consequently, the repeatability of the path followed by the joint throughout its prescribed trajectory is as important as the repeatability of the joint at reaching positions making up its trajectory, particularly when joint contact occurs.

  14. Working Partnerships: A Joint Venture in Vocational Education.

    ERIC Educational Resources Information Center

    Campbell, Clifton P.

    Joint Ventures in Vocational Education projects link participating businesses with secondary vocational programs in a cooperative relationship. These are voluntary arrangements between vocational programs and a public or private sector agency that combine the energies and resources of the partners to enrich various aspects of the vocational…

  15. Scope of Work for the Joint Mk 38 Warhead/Atlas F Missile Firing Program

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Not Available

    1964-11-16

    List of principles & responsibilities of this Sandia/AEC & Air Force joint program for verifying the functionability of the W28/AGM-28A/B-52 weapon system under actual routine handling, assembling, and firing during a stockpile-to-target sequence.

  16. Scope of Work for Joint W28/AGM-28A Missile Firing Program (Phase 1)

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Draper, E.H.

    1965-04-16

    List of principles & responsibilities of this Sandia/AEC & Air Force joint program for verifying the functionability of the W28/AGM-28A/B-52 weapon system under actual routine handling, assembling, and firing during a stockpile-to-target sequence.

  17. 78 FR 55338 - ITS Joint Program Office; Notice of Meeting

    Federal Register 2010, 2011, 2012, 2013, 2014

    2013-09-10

    ... DEPARTMENT OF TRANSPORTATION Research and Innovative Technology Administration ITS Joint Program Office; Notice of Meeting AGENCY: Research and Innovative Technology Administration (RITA), U.S. Department of Transportation. ACTION: Notice. This notice announces the cancellation of the Connected Vehicle Planning and Policy Stakeholder Meeting. Th...

  18. 75 FR 56651 - ITS Joint Program Office; Trucking Industry Mobility & Technology Coalition Annual Meeting

    Federal Register 2010, 2011, 2012, 2013, 2014

    2010-09-16

    ...: Beating Gridlock with a Smart Grid; U.S. DOT Truck Technology Initiatives; and State and Federal... DEPARTMENT OF TRANSPORTATION ITS Joint Program Office; Trucking Industry Mobility & Technology Coalition Annual Meeting AGENCY: Research and Innovative Technology Administration, U.S. Department of...

  19. JPSS-1 Prelaunch News Conference

    NASA Image and Video Library

    2017-11-12

    JPSS-1 Prelaunch News Conference at Vandenberg Air Force Base hosted by Tori Mclendon, with Steve Volz, NOAA Satellite and Information Service, Greg Mandt, Director, NOAA Joint Polar Satellite Systems Program, Sandra Smalley, NASA Joint Agency Satellite Division, Omar Baez, Launch Manager, NASA Launch Services Program, Scott Messer, Program Manager for NASA Missions, United Launch Alliance, and Ross Malugani, Launch Weather Officer, VAFB 30th Space Wing.

  20. 34 CFR 400.5 - Under what conditions may funds under the Act be used for the joint funding of programs?

    Code of Federal Regulations, 2010 CFR

    2010-07-01

    ... VOCATIONAL AND APPLIED TECHNOLOGY EDUCATION PROGRAMS-GENERAL PROVISIONS § 400.5 Under what conditions may... 34 Education 3 2010-07-01 2010-07-01 false Under what conditions may funds under the Act be used for the joint funding of programs? 400.5 Section 400.5 Education Regulations of the Offices of the...

  1. 2007 Joint Chemical Biological, Radiological and Nuclear (CBRN) Conference and Exhibition - Combating Weapons of Mass Destruction

    DTIC Science & Technology

    2007-06-27

    Selected CB Defense Systems SHAPESENSE Joint Warning and Reporting Network JSLIST CB Protected Shelter Joint Vaccine Acquisition Program Joint Effects...military can operate in any environment, unconstrained by chemical or biological weapons. 21 SHIELD SUSTAIN Selected CB Defense Systems SHAPESENSE Joint...28070625_JCBRN_Conference_Reeves UNCLASSIFIED Decontamination Vision Strippable Barriers Self-Decontaminating Fabrics/Coatings Reduce Logistics Burden

  2. Modeling the Potential Economic Impact of the Medicare Comprehensive Care for Joint Replacement Episode-Based Payment Model.

    PubMed

    Maniya, Omar Z; Mather, Richard C; Attarian, David E; Mistry, Bipin; Chopra, Aneesh; Strickland, Matt; Schulman, Kevin A

    2017-11-01

    The Medicare program has initiated Comprehensive Care for Joint Replacement (CJR), a bundled payment mandate for lower extremity joint replacements. We sought to determine the degree to which hospitals will invest in care redesign in response to CJR, and to project its economic impacts. We defined 4 potential hospital management strategies to address CJR: no action, light care management, heavy care management, and heavy care management with contracting. For each of 798 hospitals included in CJR, we used hospital-specific volume, cost, and quality data to determine the hospital's economically dominant strategy. We aggregated data to assess the percentage of hospitals pursuing each strategy; savings to the health care system; and costs and percentages of CJR-derived revenues gained or lost for Medicare, hospitals, and postacute care facilities. In the model, 83.1% of hospitals (range 55.0%-100.0%) were expected to take no action in response to CJR, and 16.1% of hospitals (range 0.0%-45.0%) were expected to pursue heavy care management with contracting. Overall, CJR is projected to reduce health care expenditures by 0.5% (range 0.0%-4.1%) or $14 million (range $0-$119 million). Medicare is expected to save 2.2% (range 2.2%-2.2%), hospitals are projected to lose 3.7% (range 4.7% loss to 3.8% gain), and postacute care facilities are expected to lose 6.5% (range 0.0%-12.8%). Hospital administrative costs are projected to increase by $63 million (range $0-$148 million). CJR is projected to have a negligible impact on total health care expenditures for lower extremity joint replacements. Further research will be required to assess the actual care management strategies adopted by CJR hospitals. Copyright © 2017 Elsevier Inc. All rights reserved.

  3. Developing a multi-joint upper limb exoskeleton robot for diagnosis, therapy, and outcome evaluation in neurorehabilitation.

    PubMed

    Ren, Yupeng; Kang, Sang Hoon; Park, Hyung-Soon; Wu, Yi-Ning; Zhang, Li-Qun

    2013-05-01

    Arm impairments in patients post stroke involve the shoulder, elbow and wrist simultaneously. It is not very clear how patients develop spasticity and reduced range of motion (ROM) at the multiple joints and the abnormal couplings among the multiple joints and the multiple degrees-of-freedom (DOF) during passive movement. It is also not clear how they lose independent control of individual joints/DOFs and coordination among the joints/DOFs during voluntary movement. An upper limb exoskeleton robot, the IntelliArm, which can control the shoulder, elbow, and wrist, was developed, aiming to support clinicians and patients with the following integrated capabilities: 1) quantitative, objective, and comprehensive multi-joint neuromechanical pre-evaluation capabilities aiding multi-joint/DOF diagnosis for individual patients; 2) strenuous and safe passive stretching of hypertonic/deformed arm for loosening up muscles/joints based on the robot-aided diagnosis; 3) (assistive/resistive) active reaching training after passive stretching for regaining/improving motor control ability; and 4) quantitative, objective, and comprehensive neuromechanical outcome evaluation at the level of individual joints/DOFs, multiple joints, and whole arm. Feasibility of the integrated capabilities was demonstrated through experiments with stroke survivors and healthy subjects.

  4. Correlation and prediction of dynamic human isolated joint strength from lean body mass

    NASA Technical Reports Server (NTRS)

    Pandya, Abhilash K.; Hasson, Scott M.; Aldridge, Ann M.; Maida, James C.; Woolford, Barbara J.

    1992-01-01

    A relationship between a person's lean body mass and the amount of maximum torque that can be produced with each isolated joint of the upper extremity was investigated. The maximum dynamic isolated joint torque (upper extremity) on 14 subjects was collected using a dynamometer multi-joint testing unit. These data were reduced to a table of coefficients of second degree polynomials, computed using a least squares regression method. All the coefficients were then organized into look-up tables, a compact and convenient storage/retrieval mechanism for the data set. Data from each joint, direction and velocity, were normalized with respect to that joint's average and merged into files (one for each curve for a particular joint). Regression was performed on each one of these files to derive a table of normalized population curve coefficients for each joint axis, direction, and velocity. In addition, a regression table which included all upper extremity joints was built which related average torque to lean body mass for an individual. These two tables are the basis of the regression model which allows the prediction of dynamic isolated joint torques from an individual's lean body mass.

  5. Joint pain undergoes a transition in accordance with signal changes of bones detected by MRI in hip osteoarthritis.

    PubMed

    Kamimura, Mikio; Nakamura, Yukio; Ikegami, Shota; Uchiyama, Shigeharu; Kato, Hiroyuki

    2013-01-01

    In this study, we aimed to investigate whether joint pain is derived from cartilage or bone alterations. We reviewed 23 hip joints of 21 patients with primary hip osteoarthritis (OA), which were classified into Kellgren-Laurence (KL) grading I to IV. Plain radiographs and magnetic resonance imaging (MRI) were obtained from all of the 23 joints. Two of the 21 patients had bilateral hip OA. Pain was assessed based on the pain scale of Denis. A Welch t test was performed for age, height, weight, body mass index, bone mineral density, and a Mann-Whitney U test was performed for KL grading. Four of 8 hip joints with pain and OA showed broad signal changes detected by MRI. Fourteen hip joints without pain, but with OA did not show broad signal changes by MRI. Collectively, MRI analyses showed that broad signal changes in OA cases without joint pain or with a slight degree of joint pain were not observed, while broad signal changes were observed in OA cases with deteriorated joint pain. Our findings suggest that hip joint pain might be associated with bone signal alterations in the hips of OA patients.

  6. Localized cervical facet joint kinematics under physiological and whiplash loading.

    PubMed

    Stemper, Brian D; Yoganandan, Narayan; Gennarelli, Thomas A; Pintar, Frank A

    2005-12-01

    Although facet joints have been implicated in the whiplash injury mechanism, no investigators have determined the degree to which joint motions in whiplash are nonphysiological. The purpose of this investigation was to quantify the correlation between facet joint and segmental motions under physiological and whiplash loading. Human cadaveric cervical spine specimens were exercise tested under physiological extension loading, and intact human head-neck complexes were exercise tested under whiplash loading to correlate the localized component motions of the C4-5 facet joint with segmental extension. Facet joint shear and distraction kinematics demonstrated a linear correlation with segmental extension under both loading modes. Facet joints responded differently to whiplash and physiological loading, with significantly increased kinematics for the same-segmental angulation. The limitations of this study include removal of superficial musculature and the limited sample size for physiological testing. The presence of increased facet joint motions indicated that synovial joint soft-tissue components (that is, synovial membrane and capsular ligament) sustain increased distortion that may subject these tissues to a greater likelihood of injury. This finding is supported by clinical investigations in which lower cervical facet joint injury resulted in similar pain patterns due to the most commonly reported whiplash symptoms.

  7. 41 CFR 101-6.212-5 - Consolidated or joint hearings.

    Code of Federal Regulations, 2010 CFR

    2010-07-01

    ..., provide for the conduct of consolidated or joint hearings, and for the application to such hearings of... 41 Public Contracts and Property Management 2 2010-07-01 2010-07-01 true Consolidated or joint...-Nondiscrimination in Programs Receiving Federal Financial Assistance § 101-6.212-5 Consolidated or joint hearings...

  8. Novel information theory-based measures for quantifying incongruence among phylogenetic trees.

    PubMed

    Salichos, Leonidas; Stamatakis, Alexandros; Rokas, Antonis

    2014-05-01

    Phylogenies inferred from different data matrices often conflict with each other necessitating the development of measures that quantify this incongruence. Here, we introduce novel measures that use information theory to quantify the degree of conflict or incongruence among all nontrivial bipartitions present in a set of trees. The first measure, internode certainty (IC), calculates the degree of certainty for a given internode by considering the frequency of the bipartition defined by the internode (internal branch) in a given set of trees jointly with that of the most prevalent conflicting bipartition in the same tree set. The second measure, IC All (ICA), calculates the degree of certainty for a given internode by considering the frequency of the bipartition defined by the internode in a given set of trees in conjunction with that of all conflicting bipartitions in the same underlying tree set. Finally, the tree certainty (TC) and TC All (TCA) measures are the sum of IC and ICA values across all internodes of a phylogeny, respectively. IC, ICA, TC, and TCA can be calculated from different types of data that contain nontrivial bipartitions, including from bootstrap replicate trees to gene trees or individual characters. Given a set of phylogenetic trees, the IC and ICA values of a given internode reflect its specific degree of incongruence, and the TC and TCA values describe the global degree of incongruence between trees in the set. All four measures are implemented and freely available in version 8.0.0 and subsequent versions of the widely used program RAxML.

  9. Modification of osteoarthritis in the guinea pig with pulsed low-intensity ultrasound treatment.

    PubMed

    Gurkan, I; Ranganathan, A; Yang, X; Horton, W E; Todman, M; Huckle, J; Pleshko, N; Spencer, R G

    2010-05-01

    The Hartley guinea pig develops articular cartilage degeneration similar to that seen in idiopathic human osteoarthritis (OA). We investigated whether the application of pulsed low-intensity ultrasound (PLIUS) to the Hartley guinea pig joint would prevent or attenuate the progression of this degenerative process. Treatment of male Hartley guinea pigs was initiated at the onset of degeneration (8 weeks of age) to assess the ability of PLIUS to prevent OA, or at a later age (12 months) to assess the degree to which PLIUS acted to attenuate the progression of established disease. PLIUS (30 mW/cm(2)) was applied to stifle joints for 20 min/day over periods ranging from 3 to 10 months, with contralateral limbs serving as controls. Joint cartilage histology was graded according to a modified Mankin scale to evaluate treatment effect. Immunohistochemical staining for interleukin-1 receptor antagonist (IL-1ra), matrix metalloproteinase (MMP)-3, MMP-13, and transforming growth factor (TGF)-beta1 was performed on the cartilage to evaluate patterns of expression of these proteins. PLIUS did not fully prevent cartilage degeneration in the prevention groups, but diminished the severity of the disease, with the treated joints showing markedly decreased surface irregularities and a much smaller degree of loss of matrix staining as compared to controls. PLIUS also attenuated disease progression in the groups with established disease, although to a somewhat lesser extent as compared to the prevention groups. Immunohistochemical staining demonstrated a markedly decreased degree of TGF-beta1 production in the PLIUS-treated joints. This indicates less active endogenous repair, consistent with the marked reduction in cartilage degradation. PLIUS exhibits the ability to attenuate the progression of cartilage degeneration in an animal model of idiopathic human OA. The effect was greater in the treatment of early, rather than established, degeneration. Published by Elsevier Ltd.

  10. The Myth of Jointness in Department of Defense Requirements and Major Acquisition Programs

    DTIC Science & Technology

    2013-06-01

    cost between the two was the result of both the need to purchase much smaller quantities for the smaller number of Defense Department aircraft, and...service requirements to reach an agreement on the joint requirements of the new aircraft. The results were borderline disastrous. The second case study...completely wrong. The TFX did not originate as a joint program but was the result of then Secretary of Defense Robert McNamara’s decision to combine

  11. Orchestration: The Movement and Vocal Behavior of Free-Ranging Norwegian Killer Whales (Orcinus Orca)

    DTIC Science & Technology

    2008-06-01

    MIT/WHOI 2008-08 Massachusetts Institute of Technology Woods Hole Oceanographic Institution 0" joint Program in Oceanography/ 0 OFTtcy Applied Ocean...Biology Paola Malanotte-Rizzoli James A. Yoder MIT Director of Joint Program WHOI Dean of Graduate Studies ORCHESTRATION: THE MOVEMENT AND VOCAL BEHAVIOR...Steve Young and Syd Lewis, have mentored and believed in me, a gift that I cherish. Stacy DeRuiter and I have been labmates since we began the Joint

  12. Multi-degree of freedom joystick for virtual reality simulation.

    PubMed

    Head, M J; Nelson, C A; Siu, K C

    2013-11-01

    A modular control interface and simulated virtual reality environment were designed and created in order to determine how the kinematic architecture of a control interface affects minimally invasive surgery training. A user is able to selectively determine the kinematic configuration of an input device (number, type and location of degrees of freedom) for a specific surgical simulation through the use of modular joints and constraint components. Furthermore, passive locking was designed and implemented through the use of inflated latex tubing around rotational joints in order to allow a user to step away from a simulation without unwanted tool motion. It is believed that these features will facilitate improved simulation of a variety of surgical procedures and, thus, improve surgical skills training.

  13. Remote manipulator system flexibility analysis program: Mission planning, mission analysis, and software formulation

    NASA Technical Reports Server (NTRS)

    Kumar, L.

    1978-01-01

    A computer program is described for calculating the flexibility coefficients as arm design changes are made for the remote manipulator system. The coefficients obtained are required as input for a second program which reduces the number of payload deployment and retrieval system simulation runs required to simulate the various remote manipulator system maneuvers. The second program calculates end effector flexibility and joint flexibility terms for the torque model of each joint for any arbitrary configurations. The listing of both programs is included in the appendix.

  14. 76 FR 2147 - UAW-Chrysler National Training Center Technology Training Joint Programs Staff, Detroit, MI; UAW...

    Federal Register 2010, 2011, 2012, 2013, 2014

    2011-01-12

    ... Technical Training Center Technology Training Joint Programs Staff, Warren, MI; Notice of Revised... facilities. The information revealed that the technical training provided (such as applied industrial... engaged in employment related to the supply of technical training services, meet the worker group...

  15. Determination of the criterion-related validity of hip joint angle test for estimating hamstring flexibility using a contemporary statistical approach.

    PubMed

    Sainz de Baranda, Pilar; Rodríguez-Iniesta, María; Ayala, Francisco; Santonja, Fernando; Cejudo, Antonio

    2014-07-01

    To examine the criterion-related validity of the horizontal hip joint angle (H-HJA) test and vertical hip joint angle (V-HJA) test for estimating hamstring flexibility measured through the passive straight-leg raise (PSLR) test using contemporary statistical measures. Validity study. Controlled laboratory environment. One hundred thirty-eight professional trampoline gymnasts (61 women and 77 men). Hamstring flexibility. Each participant performed 2 trials of H-HJA, V-HJA, and PSLR tests in a randomized order. The criterion-related validity of H-HJA and V-HJA tests was measured through the estimation equation, typical error of the estimate (TEEST), validity correlation (β), and their respective confidence limits. The findings from this study suggest that although H-HJA and V-HJA tests showed moderate to high validity scores for estimating hamstring flexibility (standardized TEEST = 0.63; β = 0.80), the TEEST statistic reported for both tests was not narrow enough for clinical purposes (H-HJA = 10.3 degrees; V-HJA = 9.5 degrees). Subsequently, the predicted likely thresholds for the true values that were generated were too wide (H-HJA = predicted value ± 13.2 degrees; V-HJA = predicted value ± 12.2 degrees). The results suggest that although the HJA test showed moderate to high validity scores for estimating hamstring flexibility, the prediction intervals between the HJA and PSLR tests are not strong enough to suggest that clinicians and sport medicine practitioners should use the HJA and PSLR tests interchangeably as gold standard measurement tools to evaluate and detect short hamstring muscle flexibility.

  16. Altered Frontal and Transverse Plane Tibiofemoral Kinematics and Patellofemoral Malalignments During Downhill Gait in Patients with Mixed Knee Osteoarthritis

    PubMed Central

    Farrokhi, Shawn; Meholic, Brad; Chuang, Wei-Neng; Gustafson, Jonathan A.; Fitzgerald, G. Kelley; Tashman, Scott

    2015-01-01

    Patients with knee osteoarthritis often present with signs of mixed tibiofemoral and patellofemoral joint disease. It has been suggested that altered frontal and transverse plane knee joint mechanics play a key role in compartment-specific patterns of knee osteoarthritis, but invivo evidence in support of this premise remains limited. Using Dynamic Stereo X-ray techniques, the aim of this study was to compare the frontal and transverse plane tibiofemoral kinematics and patellofemoral malalignments during the loading response phase of downhill gait in three groups of older adults: patients with medial tibiofemoral compartment and coexisting patellofemoral osteoarthritis (n=11); patients with lateral tibiofemoral compartment and coexisting patellofemoral osteoarthritis (n=10); and an osteoarthritis-free control group (n=22). Patients with lateral compartment osteoarthritis walked with greater and increasing degrees of tibiofemoral abduction compared to the medial compartment osteoarthritis and the control groups who walked with increasing degrees of tibiofemoral adduction. Additionally, the medial and lateral compartment osteoarthritis groups demonstrated reduced degrees of tibiofemoral internal rotation compared to the control group. Both medial and lateral compartment osteoarthritis groups also walked with increasing degrees of lateral patella tilt and medial patella translation during the loading response phase of downhill gait. Our findings suggest that despite the differences in frontal and transverse plane tibiofemoral kinematics between patients with medial and lateral compartment osteoarthritis, the malalignments of their arthritic patellofemoral joint appears to be similar. Further research is needed to determine if these kinematic variations are relevant targets for interventions to reduce pain and disease progression in patients with mixed disease. PMID:26087880

  17. Cervical spine joint hypermobility: a possible predisposing factor for new daily persistent headache.

    PubMed

    Rozen, T D; Roth, J M; Denenberg, N

    2006-10-01

    The objective of this study was to suggest that joint hypermobility (specifically of the cervical spine) is a predisposing factor for the development of new daily persistent headache (NDPH). Twelve individuals (10 female, 2 male) with primary NDPH were evaluated by one of two physical therapists. Each patient was tested for active cervical range of motion and for the presence of excessive intersegmental vertebral motion in the cervical spine. All patients were screened utilizing the Beighton score, which determines degree of systemic hypermobility. Eleven of the 12 NDPH patients were found to have cervical spine joint hypermobility. Ten of the 12 NDPH patients had evidence of widespread joint hypermobility with the Beighton score. Based on our findings we suggest that joint hypermobility, specifically of the cervical spine, may be a predisposing factor for the development of NDPH.

  18. Interacting epidemics on overlay networks

    NASA Astrophysics Data System (ADS)

    Funk, Sebastian; Jansen, Vincent A. A.

    2010-03-01

    The interaction between multiple pathogens spreading on networks connecting a given set of nodes presents an ongoing theoretical challenge. Here, we aim to understand such interactions by studying bond percolation of two different processes on overlay networks of arbitrary joint degree distribution. We find that an outbreak of a first pathogen providing immunity to another one spreading subsequently on a second network connecting the same set of nodes does so most effectively if the degrees on the two networks are positively correlated. In that case, the protection is stronger the more heterogeneous the degree distributions of the two networks are. If, on the other hand, the degrees are uncorrelated or negatively correlated, increasing heterogeneity reduces the potential of the first process to prevent the second one from reaching epidemic proportions. We generalize these results to cases where the edges of the two networks overlap to arbitrary amount, or where the immunity granted is only partial. If both processes grant immunity to each other, we find a wide range of possible situations of coexistence or mutual exclusion, depending on the joint degree distribution of the underlying networks and the amount of immunity granted mutually. These results generalize the concept of a coexistence threshold and illustrate the impact of large-scale network structure on the interaction between multiple spreading agents.

  19. Building Bridges to Diversity in Graduate Physics & Astronomy: The Fisk-Vanderbilt Masters-to-PhD Bridge Program

    NASA Astrophysics Data System (ADS)

    Stassun, Keivan G.

    2006-12-01

    We describe the Fisk-Vanderbilt Masters-to-PhD Bridge program as a successful model for effective partnerships with minority-serving institutions toward significantly broadening the participation of underrepresented groups in the physical sciences. The program couples targeted recruitment with active retention strategies, and is built upon a clearly defined structure that is flexible enough to address individual student needs while maintaining clearly communicated baseline standards for student performance. A key precept of the program’s philosophy is to eliminate passivity in student mentoring; students are deliberately groomed to successfully transition into the PhD program through active involvement in research experiences with future PhD advisers, coursework that demonstrates competency in core PhD subject areas, and frequent interactions with joint mentoring committees. This approach allows student progress and performance to be monitored and evaluated in a more holistic manner than usually afforded by limited metrics such as standardized tests. Since its inception in 2004, the program has attracted a total of 18 underrepresented students, with a retention rate of 90%. Recent research indicates that minority students are nearly twice as likely as non-minority students to seek a Masters degree en route to the PhD. In essence, the Bridge program described here builds upon this increasingly important pathway, with a dedicated mentoring process designed to ensure that the Masters-to-PhD transition is a successful one.

  20. Policing in Afghanistan-Reform that Respects Tradition: Need for a Strategic Shift

    DTIC Science & Technology

    2010-05-01

    Simpkins Log: H 10-000547 Joint Advanced Warfighting Program Approved for public release; distribution is unlimited. About This Publication This work...iii Preface The Department of Defense established the Joint Advanced Warfighting Program (JAWP) in 1998 under the senior sponsorship of the Under...2 Greg Jaffe, “ Program Aims to Rebuild Afghan Police Force, Repair its Image,” Washington Post (12 March 2010), 8, www.washingtonpost.com

  1. Joint Task Force on Undergraduate Physics Programs

    NASA Astrophysics Data System (ADS)

    This session will focus on the guidelines and recommendations being developed by the APS/AAPT Joint Task Force on Undergraduate Physics Programs. J-TUPP is studying how undergraduate physics programs might better prepare physics majors for diverse careers. The guidelines and recommendations will focus on curricular content, flexible tracks, pedagogical methods, research experiences and internships, the development of professional skills, and enhanced advising and mentoring for all physics majors.

  2. Benefits of CMM-Based Software Process Improvement: Initial Results

    DTIC Science & Technology

    1994-08-01

    Institute Carnegie Mellon University Pittsburgh, Pennsylvania 15213 This report was prepar the SEI Joint Program Office HQ ESC/ENS 5 Eglin Street Hanscom AFB...Miller, Lt Col, USAF SEI Joint Program Office This work is sponsored by the U.S. Department of Defense. Copyright 0 1994 by Carnegie Mellon University...categories: descriptive information about the organizations, information about their process improvement and measurement programs , and data about the

  3. Joint Ventures: A New Agenda for Education.

    ERIC Educational Resources Information Center

    Griffin, Dean

    1989-01-01

    The author states that vocational education can join in partnerships with industry, labor, and government to contribute to economic development. Examples of current programs are included. The "joint venture" concept is explained and ideas for forming joint ventures are shared. (CH)

  4. United States Military Posture for FY 1987

    DTIC Science & Technology

    1987-01-01

    monitor, plan, and execute mobilization, deployment, employment, and sustainment during peace, crisis, and war. JOPES requirements are being developed ...positioning 68 Deployment Management 69 VI Sustainability 69 Ammunition 69 Petroleum, Oils, and Lubricants 69 Military Construction 69...Commanders in Chiefs 73 Joint Doctrine 73 Combined Doctrine 74 Joint Force Development Process 74 Joint Resource Assessment 75 Joint Program

  5. Motion systems providing three or four degrees of freedom

    NASA Technical Reports Server (NTRS)

    Chou, Richard C. (Inventor)

    1982-01-01

    A motion system is provided by a platform generally parallel to a base and connected thereto by a column and powered and controlled extensible members, at least three of which are connected between distributed points around the column. In a three degree of freedom device, the column is conical, rigidly supported at its base with a universal joint at its top. The points of attachment define triangles in the base and in the platform surrounding the column with one extensible member connected between each. In the four degree of freedom version, the column is modified by making it effectively a column which is pivoted or guided at the base or contains an extensible member, preferably retains its triangular shape and its universal joint connection to the platform at its apex. For stability four powered and controlled extensible members are provided between points in the base and platform distributed around the column, a preferred pattern of arrangement being a square with the column at the center.

  6. Multimodal Cues in the Socialization of Joint Attention in Young Children with Varying Degrees of Vision: Getting the POINT Even When You Can't See It

    ERIC Educational Resources Information Center

    Rickard, Carolyn

    2013-01-01

    Research on joint attention and language learning has focused primarily on cues requiring visual access. However, this narrow focus cannot account for the emergence of language among some congenitally blind children who develop language on the same developmental timescale as their sighted peers. Findings from this longitudinal, retrospective study…

  7. 12th Annual Small Business Conference

    DTIC Science & Technology

    2008-11-13

    Joint Combat support systems. CoL myers graduated from marshall University in huntington, west virginia with a bachelors degree in Communications. he...as director and Principal assistant responsible for Contracting (ParC) for the Us army Contracting Command’s rock island Contracting Center (riCC...sustainment Command (asC) and the Joint munitions & Lethality Life Cycle management Command (Jm&L-LCmC) – rock island . in this position, he advises

  8. Finite element analysis of a deployable space structure

    NASA Technical Reports Server (NTRS)

    Hutton, D. V.

    1982-01-01

    To assess the dynamic characteristics of a deployable space truss, a finite element model of the Scientific Applications Space Platform (SASP) truss has been formulated. The model incorporates all additional degrees of freedom associated with the pin-jointed members. Comparison of results with SPAR models of the truss show that the joints of the deployable truss significantly affect the vibrational modes of the structure only if the truss is relatively short.

  9. Early Postoperative Results of Percutaneous Needle Fasciotomy in 451 Patients with Dupuytren Disease.

    PubMed

    Molenkamp, Sanne; Schouten, Tanneke A M; Broekstra, Dieuwke C; Werker, Paul M N; Moolenburgh, J Daniel

    2017-06-01

    Percutaneous needle fasciotomy is a minimally invasive treatment modality for Dupuytren disease. In this study, the authors analyzed the efficacy and complication rate of percutaneous needle fasciotomy using a statistical method that takes the multilevel structure of data, regarding multiple measurements from the same patient, into account. The data of 470 treated rays from 451 patients with Dupuytren disease that underwent percutaneous needle fasciotomy were analyzed retrospectively. The authors described the early postoperative results of percutaneous needle fasciotomy and applied linear mixed models to compare mean correction of passive extension deficit between joints and efficacy of primary versus secondary percutaneous needle fasciotomy. Mean preoperative passive extension deficits at the metacarpophalangeal, proximal interphalangeal, and distal interphalangeal joints were 37, 40, and 31 degrees, respectively. Mean preoperative total passive extension deficit was 54 degrees. Results were excellent, with a mean total passive extension deficit correction of 85 percent. Percutaneous needle fasciotomy was most effective for metacarpophalangeal joints and less effective for proximal interphalangeal and distal interphalangeal joints. Secondary percutaneous needle fasciotomy was as effective as primary percutaneous needle fasciotomy. Complications were rare and mostly minor. The results of this study confirm that percutaneous needle fasciotomy is an effective and safe treatment modality for patients with mild to moderate disease who prefer a minimally invasive procedure. Therapeutic, IV.

  10. The kinematic recovery process of rhesus monkeys after spinal cord injury.

    PubMed

    Wei, Rui-Han; Zhao, Can; Rao, Jia-Sheng; Zhao, Wen; Zhou, Xia; Tian, Peng-Yu; Song, Wei; Ji, Run; Zhang, Ai-Feng; Yang, Zhao-Yang; Li, Xiao-Guang

    2018-05-16

    After incomplete spinal cord injury (SCI), neural circuits may be plastically reconstructed to some degree, resulting in extensive functional locomotor recovery. The present study aimed to observe the post-SCI locomotor recovery of rhesus monkey hindlimbs and compare the recovery degrees of different hindlimb parts, thus revealing the recovery process of locomotor function. Four rhesus monkeys were chosen for thoracic hemisection injury. The hindlimb locomotor performance of these animals was recorded before surgery, as well as 6 and 12 weeks post-lesion. Via principal component analysis, the relevant parameters of the limb endpoint, pelvis, hindlimb segments, and joints were processed and analyzed. Twelve weeks after surgery, partial kinematic recovery was observed at the limb endpoint, shank, foot, and knee joints, and the locomotor performance of the ankle joint even recovered to the pre-lesion level; the elevation angle of the thigh and hip joints showed no obvious recovery. Generally, different parts of a monkey hindlimb had different spontaneous recovery processes; specifically, the closer the part was to the distal end, the more extensive was the locomotor function recovery. Therefore, we speculate that locomotor recovery may be attributed to plastic reconstruction of the motor circuits that are mainly composed of corticospinal tract. This would help to further understand the plasticity of motor circuits after spinal cord injury.

  11. [Case report of a patient with ochronosis and arthroplasty of the hip and both knees].

    PubMed

    Moslavac, Aleksandra; Moslavac, Sasa; Cop, Renata

    2003-01-01

    Alkaptonuria is a rare hereditary metabolic disorder characterised by absence of the enzyme homogentisic acid oxidase. As a result of this defect homogentisic acid accumulates and is excreted in the urine. The term ochronosis is used to describe bluish-black pigmentation of connective tissue. Ochronotic arthropathy results from the pigmented deposits in the joints of the appendicular and axial skeleton. Findings simulate those of uncomplicated degenerative joint disease, with effusion, articular space narrowing, and bony sclerosis. Our patient is a 70-year old male with ochronotic arthropathy. He has typical ears and sclera discoloration, and had arthroplasty of knees 7 and 4 years ago, respectively. In year 2002, he had undergone total right hip arthroplasty and has been admitted for rehabilitation 14th postoperative day. Individually designed rehabilitation regimen included kinesitherapy, hydrokinesitherapy, and ambulation training with gradual increase in weight bearing exercises and electro-analgesia of associated low back pain. In course of rehabilitation our patient improved his endurance with satisfying range of motion of right hip (flexion 90 degrees, abduction 40 degrees) and strength of hip and thigh musculature. The patient was able to walk with crutches without limitation. We conclude that joint destruction followed by painful locomotion due to ochronotic arthropathy is best treated by total joint arthroplasty, as described in our patient.

  12. A Multiple Degree of Freedom Lower Extremity Isometric Device to Simultaneously Quantify Hip, Knee and Ankle Torques

    PubMed Central

    Sánchez, Natalia; Acosta, Ana Maria; Stienen, Arno H.A.

    2015-01-01

    Characterization of the joint torque coupling strategies used in the lower extremity to generate maximal and submaximal levels of torque at either the hip, knee or ankle is lacking. Currently, there are no available isometric devices that quantify all concurrent joint torques in the hip, knee and ankle of a single leg during maximum voluntary torque generation. Thus, joint-torque coupling strategies in the hip, knee and concurrent torques at ankle and/or coupling patterns at the hip and knee driven by the ankle have yet to be quantified. This manuscript describes the design, implementation and validation of a multiple degree of freedom, lower extremity isometric device (the MultiLEIT) that accurately quantifies simultaneous torques at the hip, knee and ankle. The system was mechanically validated and then implemented with two healthy control individuals and two post-stroke individuals to test usability and patient acceptance. Data indicated different joint torque coupling strategies used by both healthy individuals. In contrast, data showed the same torque coupling patterns in both post-stroke individuals, comparable to those described in the clinic. Successful implementation of the MultiLEIT can contribute to the understanding of the underlying mechanisms responsible for abnormal movement patterns and aid in the design of therapeutic interventions. PMID:25163064

  13. Repair of acute injuries of the lateral ligament complex of the ankle by suture anchors

    PubMed Central

    Liu, Xiang-Fei; Fang, Yang; Cao, Zhong-Hua; Li, Guang-Feng; Yang, Guo-Qing

    2015-01-01

    Objective: The objective of this study was to investigate the clinical curative effect of stage I repair of acute injuries of the lateral ligament complex of the ankle by the application of suture anchors. Methods: We retrospectively analyzed 18 cases of III degree acute injuries of the lateral ligament complex of the ankle. Results: There were statistically significant differences in preoperative and last follow-up VAS pain scores and AOFAS ankle hind-foot function scores. The X-ray talus displacement values in the anterior drawer test and pressure anteroposterior X-ray talar tilt in the ankle talar tilt test also showed statistically significant differences. Complications occurred in 2 patients, incision surface infection in one, and postoperative lateral dorsal skin numbness in one. All these cases were cured after symptomatic treatment. At the last follow-up all patients’ ankle joint activity recovered to their preinjury function levels. Conclusion: The application of suture anchors for small incision stage I repair of the lateral collateral ligament of ankle joint degree III injury, can effectively restored the stability of ankle joint, and prevent the occurrence of chronic ankle instability complications. It is effective and feasible for the treatment of ankle joint lateral collateral ligament injuries. PMID:26885144

  14. FAA/NASA Joint University Program for Air Transportation Research, 1992-1993

    NASA Technical Reports Server (NTRS)

    Morrell, Frederick R. (Compiler)

    1994-01-01

    The research conducted during the academic year 1992-1993 under the FAA/NASA sponsored Joint University Program for Air Transportation Research is summarized. The year end review was held at Ohio University, Athens, Ohio, 17-18 June 1993. The Joint University Program is a coordinated set of three grants sponsored by the Federal Aviation Administration and NASA Langley Research Center, one each with the Massachusetts Institute of Technology, Ohio University, and Princeton University. Completed works, status reports, and annotated bibliographies are presented for research topics, which include navigation, guidance, and control theory and practice, aircraft performance, human factors and air traffic management. An overview of the year's activities for each university is also presented.

  15. Adaptation of an Evidence-Based Arthritis Program for Breast Cancer Survivors on Aromatase Inhibitor Therapy Who Experience Joint Pain

    PubMed Central

    Callahan, Leigh F.; Rini, Christine; Altpeter, Mary; Hackney, Betsy; Schecher, Arielle; Wilson, Anne; Muss, Hyman B.

    2015-01-01

    Adding aromatase inhibitors (AIs) to adjuvant treatment of postmenopausal women with hormone-receptor–positive breast cancer significantly reduces cancer recurrence. A common side effect of AIs is noninflammatory joint pain and stiffness (arthralgia) similar to arthritis symptoms. An evidence-based walking program developed by the Arthritis Foundation — Walk With Ease (WWE) — reduces arthritis-related joint symptoms. We hypothesized that WWE may also reduce AI-associated arthralgia. However, the potential for different barriers and facilitators to physical activity for these 2 patient populations suggested a need to adapt WWE before testing it with breast cancer survivors. We conducted qualitative research with 46 breast cancer survivors to explore program modification and inform the development of materials for an adapted program (Walk With Ease-Breast Cancer). Our process parallels the National Cancer Institute’s Research-Tested Intervention Programs (RTIPs) guidelines for adapting evidence-based programs for cancer populations. Findings resulted in a customized 8-page brochure to supplement existing WWE materials. PMID:26068412

  16. Design/Analysis of the JWST ISIM Bonded Joints for Survivability at Cryogenic Temperatures

    NASA Technical Reports Server (NTRS)

    Bartoszyk, Andrew; Johnston, John; Kaprielian, Charles; Kuhn, Jonathan; Kunt, Cengiz; Rodini,Benjamin; Young, Daniel

    1990-01-01

    A major design and analysis challenge for the JWST ISIM structure is thermal survivability of metal/composite bonded joints below the cryogenic temperature of 30K (-405 F). Current bonded joint concepts include internal invar plug fittings, external saddle titanium/invar fittings and composite gusset/clip joints all bonded to M55J/954-6 and T300/954-6 hybrid composite tubes (75mm square). Analytical experience and design work done on metal/composite bonded joints at temperatures below that of liquid nitrogen are limited and important analysis tools, material properties, and failure criteria for composites at cryogenic temperatures are sparse in the literature. Increasing this challenge is the difficulty in testing for these required tools and properties at cryogenic temperatures. To gain confidence in analyzing and designing the ISIM joints, a comprehensive joint development test program has been planned and is currently running. The test program is designed to produce required analytical tools and develop a composite failure criterion for bonded joint strengths at cryogenic temperatures. Finite element analysis is used to design simple test coupons that simulate anticipated stress states in the flight joints; subsequently the test results are used to correlate the analysis technique for the final design of the bonded joints. In this work, we present an overview of the analysis and test methodology, current results, and working joint designs based on developed techniques and properties.

  17. A model-based approach to stabilizing crutch supported paraplegic standing by artificial hip joint stiffness.

    PubMed

    van der Spek, Jaap H; Veltink, Peter H; Hermens, Hermie J; Koopman, Bart F J M; Boom, Herman B K

    2003-12-01

    The prerequisites for stable crutch supported standing were analyzed in this paper. For this purpose, a biomechanical model of crutch supported paraplegic stance was developed assuming the patient was standing with extended knees. When using crutches during stance, the crutches will put a position constraint on the shoulder, thus reducing the number of degrees of freedom. Additional hip-joint stiffness was applied to stabilize the hip joint and, therefore, to stabilize stance. The required hip-joint stiffness for changing crutch placement and hip-joint offset angle was studied under static and dynamic conditions. Modeling results indicate that, by using additional hip-joint stiffness, stable crutch supported paraplegic standing can be achieved, both under static as well as dynamic situations. The static equilibrium postures and the stability under perturbations were calculated to be dependent on crutch placement and stiffness applied. However, postures in which the hip joint was in extension (C postures) appeared to the most stable postures. Applying at least 60 N x m/rad hip-joint stiffness gave stable equilibrium postures in all cases. Choosing appropriate hip-joint offset angles, the static equilibrium postures changed to more erect postures, without causing instability or excessive arm forces to occur.

  18. The effect of instability training on knee joint proprioception and core strength.

    PubMed

    Cuğ, Mutlu; Ak, Emre; Ozdemir, Recep Ali; Korkusuz, Feza; Behm, David G

    2012-01-01

    Although there are many studies demonstrating increased trunk activation under unstable conditions, it is not known whether this increased activation would translate into meaningful trunk strength with a prolonged training program. Additionally, while balance-training programs have been shown to improve stability, their effect on specific joint proprioception is not clear. Thus the objective of this study was to examine training adaptations associated with a 10-week instability-training program. Participants were tested pre- and post-training for trunk extension and flexion strength and knee proprioception. Forty-three participants participated in either a 10-week (3 days per week) instability-training program using Swiss balls and body weight as resistance or a control group (n = 17). The trained group increased (p < 0. 05) trunk extension peak torque/body weight (23.6%) and total work output (20.1%) from pre- to post-training while the control group decreased by 6.8% and 6.7% respectively. The exercise group increased their trunk flexion peak torque/body weight ratios by 18.1% while the control group decreased by 0.4%. Knee proprioception (combined right and left joint repositioning) improved 44.7% from pre- to post-training (p = 0.0006) and persisted (21.5%) for 9 months post-training. In addition there was a side interaction with the position sense of the right knee at 9 months showing 32.1% (p = 0.03) less deviation from the reference angle than the right knee during pre-testing. An instability-training program using Swiss balls with body weight as resistance can provide prolonged improvements in joint proprioception and core strength in previously untrained individuals performing this novel training stress which would contribute to general health. Key pointsAlthough traditional free weight resistance exercises have been recommended as most beneficial for improving strength and power in athletes (Behm et al., 2010b), an IT program using Swiss balls and body weight as a resistance may provide an alternative starting point for the sedentary untrained population.As it is well documented that force or strength is decreased when unbalanced (Behm et al., 2010b) and balance-training programs improve balance (Behm and Kean 2006), this type of instability RT program can provide significant adaptations to improve trunk strength especially with the untrained.This type of training should also be incorporated into a new program as the improvements in joint proprioception may help protect from joint injuries over a protracted period.The finding that improved joint proprioception persists for months after training should be emphasized to those individuals whose training is regularly or inconsistently interrupted.

  19. Fluorescence optical imaging and 3T-MRI for detection of synovitis in patients with rheumatoid arthritis in comparison to a composite standard of reference.

    PubMed

    Thuermel, Klaus; Neumann, Jan; Jungmann, Pia M; Schäffeler, Christoph; Waldt, Simone; Heinze, Alexander; Beckmann, Alexander; Hauser, Christine; Hasenau, Anna-Lena; Wildgruber, Moritz; Clotten, Sigrun; Sievert, Matti; Haller, Bernhard; Woertler, Klaus; Harasser, Norbert; Rummeny, Ernst J; Meier, Reinhard

    2017-05-01

    To address whether Indocyanine Green (ICG) enhanced fluorescence optical imaging (FOI) is more sensitive than magnetic resonance imaging (MRI) in the detection of synovitis of the wrist and finger joints in rheumatoid arthritis and to analyze the performance of FOI depending on the grade of synovitis. Twenty patients with highly active rheumatoid arthritis (mean DAS28-ESR 5.25±1.0) and thirteen healthy volunteers underwent clinical examination, FOI and contrast-enhanced 3T-MRI. Joints were rated by three independent readers semiquantitatively (grade 0-3: no, low, moderate and high grade synovitis) and compared to a semiquantitative composite standard of reference (cSOR, grade 0-3) that incorporated clinical parameters, FOI and MRI results. 2.868 evaluations in 956 joints were performed. FOI had an overall sensitivity of 57.3% and a specificity of 92.1%, whereas MRI had a sensitivity of 89.2% and a specificity of 92.6%. The sensitivity of FOI increased with the degree of synovitis to 65.0% for moderate and severe synovitis (specificity 88.1%) and 76,3% for severe synovitis (specificity 80.5%). The performance of FOI decreased with the degree of synovitis with false negative results predominantly for mild (156/343, 45.5%) and moderate (160/343, 46.6%) synovitis and false positive FOI evaluations predominantly based on weak (grade 1) signals (133/163, 81,6%). FOI has a lower sensitivity than 3T-MRI in the detection of synovitis of the hand and finger joints. The diagnostic performance of FOI decreases with the degree of synovitis and with the strength of FOI signals. Copyright © 2017 Elsevier B.V. All rights reserved.

  20. An evaluation of pharmacology curricula in Australian science and health-related degree programs.

    PubMed

    Lloyd, Hilary; Hinton, Tina; Bullock, Shane; Babey, Anna-Marie; Davis, Elizabeth; Fernandes, Lynette; Hart, Joanne; Musgrave, Ian; Ziogas, James

    2013-11-19

    Pharmacology is a biomedical discipline taught in basic science and professional degree programs. In order to provide information that would facilitate pharmacology curricula to be refined and developed, and approaches to teaching to be updated, a national survey was undertaken in Australia that investigated pharmacology course content, teaching and summative assessment methods. Twenty-two institutions participated in a purpose-built online questionnaire, which enabled an evaluation of 147 courses taught in 10 different degrees. To enable comparison, degrees were grouped into four major degree programs, namely science, pharmacy, medicine and nursing. The pharmacology content was then classified into 16 lecture themes, with 2-21 lecture topics identified per theme. The resultant data were analysed for similarities and differences in pharmacology curricula across the degree programs. While all lecture themes were taught across degree programs, curriculum content differed with respect to the breadth and hours of coverage. Overall, lecture themes were taught most broadly in medicine and with greatest coverage in pharmacy. Reflecting a more traditional approach, lectures were a dominant teaching method (at least 90% of courses). Sixty-three percent of science courses provided practical classes but such sessions occurred much less frequently in other degree programs, while tutorials were much more common in pharmacy degree programs (70%). Notably, problem-based learning was common across medical programs. Considerable diversity was found in the types of summative assessment tasks employed. In science courses the most common form of in-semester assessment was practical reports, whereas in other programs pen-and-paper quizzes predominated. End-of-semester assessment contributed 50-80% to overall assessment across degree programs. The similarity in lecture themes taught across the four different degree programs shows that common knowledge- and competency-based learning outcomes can be defined for pharmacology. The authors contend that it is the differences in breadth and coverage of material for each lecture theme, and the differing teaching modes and assessment that characterise particular degree programs. Adoption of pharmacology knowledge-based learning outcomes that could be tailored to suit individual degree programs would better facilitate the sharing of expertise and teaching practice than the current model where pharmacology curricula are degree-specific.

  1. 76 FR 28040 - Medicare and Medicaid Programs; Application by the Joint Commission for Continued Deeming...

    Federal Register 2010, 2011, 2012, 2013, 2014

    2011-05-13

    ... furnish information for use in enforcement activities; monitoring procedures for provider entities found... agreements are at 42 CFR part 489 and those pertaining to activities relating to the survey and certification... for monitoring CAHs found out of compliance with the Joint Commission's program requirements. These...

  2. Assessment of the Accounting and Joint Accounting/Computer Information Systems Programs.

    ERIC Educational Resources Information Center

    Appiah, John; Cernigliaro, James; Davis, Jeffrey; Gordon, Millicent; Richards, Yves; Santamaria, Fernando; Siegel, Annette; Lytle, Namy; Wharton, Patrick

    This document presents City University of New York LaGuardia Community College's Department of Accounting and Managerial Studies assessment of its accounting and joint accounting/computer information systems programs report, and includes the following items: (1) description of the mission and goals of the Department of Accounting and Managerial…

  3. 45 CFR 1355.35 - Program improvement plans.

    Code of Federal Regulations, 2013 CFR

    2013-10-01

    ... developed jointly by title IV-E agency and Federal staff in consultation with the review team; (ii) Identify... how the action steps in the plan build on and make progress over prior program improvement plans; (vii... evaluated jointly by the title IV-E agency and ACF, in collaboration with other members of the review team...

  4. Activity Analysis and Cost Analysis in Medical Schools.

    ERIC Educational Resources Information Center

    Koehler, John E.; Slighton, Robert L.

    There is no unique answer to the question of what an ongoing program costs in medical schools. The estimates of program costs generated by classical methods of cost accounting are unsatisfactory because such accounting cannot deal with the joint production or joint cost problem. Activity analysis models aim at calculating the impact of alternative…

  5. Finding the Worker: Adult Education and Workers' Education

    ERIC Educational Resources Information Center

    Rose, Amy D.; Jeris, Laurel H.

    2011-01-01

    This article looks at how administrators and teachers who consider themselves adult educators but who find themselves in a union environment come to think about their role and their context. This article is based on research conducted with administrators and teachers working in a joint union-employer sponsored program. Joint programs were…

  6. Joint Advertising Market Research & Studies (JAMRS)

    Science.gov Websites

    Market Research & Studies Marketing Communications Recruiting Database Affiliations WELCOME TO JOINT joint marketing communications and market research and studies. One of JAMRS' objectives is to explore reported to Congress. Our marketing communications programs help increase awareness and broaden people's

  7. Joint Chance-Constrained Dynamic Programming

    NASA Technical Reports Server (NTRS)

    Ono, Masahiro; Kuwata, Yoshiaki; Balaram, J. Bob

    2012-01-01

    This paper presents a novel dynamic programming algorithm with a joint chance constraint, which explicitly bounds the risk of failure in order to maintain the state within a specified feasible region. A joint chance constraint cannot be handled by existing constrained dynamic programming approaches since their application is limited to constraints in the same form as the cost function, that is, an expectation over a sum of one-stage costs. We overcome this challenge by reformulating the joint chance constraint into a constraint on an expectation over a sum of indicator functions, which can be incorporated into the cost function by dualizing the optimization problem. As a result, the primal variables can be optimized by a standard dynamic programming, while the dual variable is optimized by a root-finding algorithm that converges exponentially. Error bounds on the primal and dual objective values are rigorously derived. We demonstrate the algorithm on a path planning problem, as well as an optimal control problem for Mars entry, descent and landing. The simulations are conducted using a real terrain data of Mars, with four million discrete states at each time step.

  8. Femoral anatomical frame: assessment of various definitions.

    PubMed

    Della Croce, U; Camomilla, V; Leardini, A; Cappozzo, A

    2003-06-01

    The reliability of the estimate of joint kinematic variables and the relevant functional interpretation are affected by the uncertainty with which bony anatomical landmarks and underlying bony segment anatomical frames are determined. When a stereo-photogrammetric system is used for in vivo studies, minimising and compensating for this uncertainty is crucial. This paper deals with the propagation of the errors associated with the location of both internal and palpable femoral anatomical landmarks to the estimation of the orientation of the femoral anatomical frame and to the knee joint angles during movement. Given eight anatomical landmarks, and the precision with which they can be identified experimentally, 12 different rules were defined for the construction of the anatomical frame and submitted to comparative assessment. Results showed that using more than three landmarks allows for more repeatable anatomical frame orientation and knee joint kinematics estimation. Novel rules are proposed that use optimization algorithms. On the average, the femoral frame orientation dispersion had a standard deviation of 2, 2.5 and 1.5 degrees for the frontal, transverse, and sagittal plane, respectively. However, a proper choice of the relevant construction rule allowed for a reduction of these inaccuracies in selected planes to 1 degrees rms. The dispersion of the knee adduction-abduction and internal-external rotation angles could also be limited to 1 degrees rms irrespective of the flexion angle value.

  9. Faying Surface Lubrication Effects on Nut Factors

    NASA Technical Reports Server (NTRS)

    Taylor, Deneen M.; Morrison, Raymond F.

    2006-01-01

    Bolted joint analysis typically is performed using nut factors derived from textbooks and procedures from program requirement documents. Joint specific testing was performed for a critical International Space Station (ISS) joint. Test results indicate that for some configurations the nut factor may be significantly different than accepted textbook values. This paper presents results of joint specific testing to aid in determining if joint specific testing should be performed to insure required preloads are obtained.

  10. Telerobotic control of the seven-degree-of-freedom CESAR manipulator

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Babcock, S.M.; Dubey, R.V.; Euler, J.A.

    1988-01-01

    The application of a computationally efficient kinematic control scheme for manipulators with redundant degrees of freedom to the unilateral telerobotic control of seven-degree-of-freedom manipulator (CESARM) at the Oak Ridge National Laboratory Center for Engineering Systems Advanced Research is presented. The kinematic control scheme uses a gradient projection optimization method, which eliminates that need to determine the generalized inverse of the Jacobian when solving for joint velocities, given Cartesian end-effector velocities. A six-degree-of-freedom (nonreplica) master controller is used. Performance indices for redundancy resolution are discussed. 5 ref., 6 figs.

  11. Kinematic performance of a six degree-of-freedom hand model (6DHand) for use in occupational biomechanics.

    PubMed

    Buczek, Frank L; Sinsel, Erik W; Gloekler, Daniel S; Wimer, Bryan M; Warren, Christopher M; Wu, John Z

    2011-06-03

    Upper extremity musculoskeletal disorders represent an important health issue across all industry sectors; as such, the need exists to develop models of the hand that provide comprehensive biomechanics during occupational tasks. Previous optical motion capture studies used a single marker on the dorsal aspect of finger joints, allowing calculation of one and two degree-of-freedom (DOF) joint angles; additional algorithms were needed to define joint centers and the palmar surface of fingers. We developed a 6DOF model (6DHand) to obtain unconstrained kinematics of finger segments, modeled as frusta of right circular cones that approximate the palmar surface. To evaluate kinematic performance, twenty subjects gripped a cylindrical handle as a surrogate for a powered hand tool. We hypothesized that accessory motions (metacarpophalangeal pronation/supination; proximal and distal interphalangeal radial/ulnar deviation and pronation/supination; all joint translations) would be small (less than 5° rotations, less than 2mm translations) if segment anatomical reference frames were aligned correctly, and skin movement artifacts were negligible. For the gripping task, 93 of 112 accessory motions were small by our definition, suggesting this 6DOF approach appropriately models joints of the fingers. Metacarpophalangeal supination was larger than expected (approximately 10°), and may be adjusted through local reference frame optimization procedures previously developed for knee kinematics in gait analysis. Proximal translations at the metacarpophalangeal joints (approximately 10mm) were explained by skin movement across the metacarpals, but would not corrupt inverse dynamics calculated for the phalanges. We assessed performance in this study; a more rigorous validation would likely require medical imaging. Published by Elsevier Ltd.

  12. Drug-food interaction counseling programs in teaching hospitals.

    PubMed

    Wix, A R; Doering, P L; Hatton, R C

    1992-04-01

    The results of a survey to characterize drug-food interaction counseling programs in teaching hospitals and solicit opinions on these programs from pharmacists and dietitians are reported. A questionnaire was mailed to the pharmacy director and the director of dietary services at teaching hospitals nationwide. The questionnaire contained 33 questions relating to hospital characteristics, drug-food interaction counseling programs, and the standard calling for such programs issued by the Joint Commission on Accreditation of Healthcare Organizations. Of 792 questionnaires mailed, 425 were returned (response rate, 53.7). A majority of the pharmacists and dietitians (51.2%) did not consider their drug-food interaction counseling program to be formal; some had no program. The pharmacy department was involved more in program development than in the daily operation of such programs. The most frequent methods of identifying patients for counseling were using lists of patients' drugs and using physicians' orders. A mean of only five drugs were targeted per program. Slightly over half the respondents rated the Joint Commission standard less effective than other standards in its ability to improve patient care. A majority of teaching hospitals did not have formal drug-food interaction counseling programs. Pharmacists and dietitians did not view these programs as greatly beneficial and did not believe that the Joint Commission has clearly delineated the requirements for meeting its standard.

  13. Proposed Accreditation Standards for Degree-Granting Correspondence Programs Offered by Accredited Institutions.

    ERIC Educational Resources Information Center

    McGraw-Hill Continuing Education Center, Washington, DC.

    A study on proposed accreditation standards grew out of a need to (1) stimulate the growth of quality correspondence degree programs; and (2) provide a policy for accreditation of correspondence degree programs so that graduates would be encouraged to pursue advanced degree programs offered elsewhere by educational institutions. The study focused…

  14. Joint hypermobility in patients with chondromalacia patellae.

    PubMed

    al-Rawi, Z; Nessan, A H

    1997-12-01

    The relationship between joint mobility and chondromalacia patellae was reported in a prospective study. A total of 115 patients with chondromalacia patellae were compared with 110 healthy individuals without chondromalacia patellae, matched for age and sex, who served as a control group. The degree of joint mobility was scored on a scale of 0-9. The number of individuals with hypermobile joints and the total mobility scores were significantly higher in patients with chondromalacia patellae when compared to the control group (P < 0.001). There were more hypermobile knees among knee joints with chondromalacia patellae when compared with the knees of the control group (P < 0.01). Chondromalacia patellae were bilateral in 57% of our patients. It occurred more frequently in the longer leg and was associated with quadriceps muscle wasting in 50% of patients. Flat feet and backache were reported significantly more often in patients compared with the control group (P < 0.05). It is concluded that hypermobility of the knee joint may be a contributing factor in the pathogenesis of chondromalacia patellae.

  15. Development of a Wearable Assist Robot for Walk Rehabilitation After Knee Arthroplasty Surgery

    NASA Astrophysics Data System (ADS)

    Terada, H.; Zhu, Y.; Horiguchi, K.; Nakamura, M.; Takahashi, R.

    In Japan, it is popular that the disease knee joints will be replaced to artificial joints by surgery. And we have to assist so many patients for walk rehabilitation. So, the wearable assist robot has been developed. This robot includes the knee motion assist mechanism and the hip joint support mechanism. Especially, the knee motion assist mechanism consists of a non-circular gear and grooved cams. This mechanism rotates and slides simultaneously, which has two degree-of-freedom. Also, the hip joint support mechanism consists of a hip brace and a ball-joint. This mechanism can avoid motion constraints which are the internal or external rotation and the adduction or abduction. Then, the control algorithm, which considers an assisting timing for the walk rehabilitation, has been proposed. A sensing system of a walk state for this control system uses a heel contacts sensor and knee and hip joint rotation angle sensors. Also, the prototype robot has been tested. And it is confirmed that the assisting system is useful.

  16. F-111E IPCS in flight

    NASA Technical Reports Server (NTRS)

    1975-01-01

    This NASA Dryden Flight Research Center photograph taken in 1975 shows the General Dynamic IPCS/F-111E Aardvark with a camouflage paint pattern. This prototype F-111E was used during the flight testing of the Integrated Propulsion Control System (IPCS). The wings of the IPCS/F-111E are swept back to near 60 degrees for supersonic flight. During the same period as F-111 TACT program, an F-111E Aardvark (#67-0115) was flown at the NASA Flight Research Center to investigate an electronic versus a conventional hydro-mechanical controlled engine. The program called integrated propulsion control system (IPCS) was a joint effort by NASA's Lewis Research Center and Flight Research Center, the Air Force's Flight Propulsion Laboratory and the Boeing, Honeywell and Pratt & Whitney companies. The left engine of the F-111E was selected for modification to an all electronic system. A Pratt & Whitney TF30-P-9 engine was modified and extensively laboratory, and ground-tested before installation into the F-111E. There were 14 IPCS flights made from 1975 through 1976. The flight demonstration program proved an engine could be controlled electronically, leading to a more efficient Digital Electronic Engine Control System flown in the F-15.

  17. Education of biomedical engineering in Taiwan.

    PubMed

    Lin, Kang-Ping; Kao, Tsair; Wang, Jia-Jung; Chen, Mei-Jung; Su, Fong-Chin

    2014-01-01

    Biomedical Engineers (BME) play an important role in medical and healthcare society. Well educational programs are important to support the healthcare systems including hospitals, long term care organizations, manufacture industries of medical devices/instrumentations/systems, and sales/services companies of medical devices/instrumentations/system. In past 30 more years, biomedical engineering society has accumulated thousands people hold a biomedical engineering degree, and work as a biomedical engineer in Taiwan. Most of BME students can be trained in biomedical engineering departments with at least one of specialties in bioelectronics, bio-information, biomaterials or biomechanics. Students are required to have internship trainings in related institutions out of campus for 320 hours before graduating. Almost all the biomedical engineering departments are certified by IEET (Institute of Engineering Education Taiwan), and met the IEET requirement in which required mathematics and fundamental engineering courses. For BMEs after graduation, Taiwanese Society of Biomedical Engineering (TSBME) provides many continue-learning programs and certificates for all members who expect to hold the certification as a professional credit in his working place. In current status, many engineering departments in university are continuously asked to provide joint programs with BME department to train much better quality students. BME is one of growing fields in Taiwan.

  18. Strength evaluation of socket joints

    NASA Technical Reports Server (NTRS)

    Rash, Larry C.

    1994-01-01

    This report documents the development of a set of equations that can be used to provide a relatively simple solution for identifying the strength of socket joints and for most cases avoid the need of more lengthy analyses. The analytical approach was verified by comparison of the contact load distributions to results obtained from a finite element analysis. The contacting surfaces for the specific joint in this analysis are in the shape of frustrums of a cone and are representative of the tapered surfaces in the socket-type joints used to join segments of model support systems for wind tunnels. The results are in the form of equations that can be used to determine the contact loads and stresses in the joint from the given geometry and externally applied loads. Equations were determined to define the bending moments and stresses along the length of the joints based on strength and materials principles. The results have also been programmed for a personal computer and a copy of the program is included.

  19. SEPs to finger joint input lack the N20-P20 response that is evoked by tactile inputs: contrast between cortical generators in areas 3b and 2 in humans.

    PubMed

    Desmedt, J E; Ozaki, I

    1991-01-01

    A method using a DC servo motor is described to produce brisk angular movements at finger interphalangeal joints in humans. Small passive flexions of 2 degrees elicited sizable somatosensory evoked potentials (SEPs) starting with a contralateral positive P34 parietal response thought to reflect activation of a radial equivalent dipole generator in area 2 which receives joint inputs. By contrast, electric stimulation of tactile (non-joint) inputs from the distal phalanx evoked the usual contralateral negative N20 reflecting a tangential equivalent dipole generator in area 3b. Finger joint inputs also evoked a precentral positivity equivalent to the P22 of motor area 4, and a large frontal negativity equivalent to N30. It is suggested that natural stimulation allows human SEP components to be differentiated in conjunction with distinct cortical somatotopic projections.

  20. BioMot exoskeleton - Towards a smart wearable robot for symbiotic human-robot interaction.

    PubMed

    Bacek, Tomislav; Moltedo, Marta; Langlois, Kevin; Prieto, Guillermo Asin; Sanchez-Villamanan, Maria Carmen; Gonzalez-Vargas, Jose; Vanderborght, Bram; Lefeber, Dirk; Moreno, Juan C

    2017-07-01

    This paper presents design of a novel modular lower-limb gait exoskeleton built within the FP7 BioMot project. Exoskeleton employs a variable stiffness actuator in all 6 joints, a directional-flexibility structure and a novel physical humanrobot interfacing, which allows it to deliver the required output while minimally constraining user's gait by providing passive degrees of freedom. Due to modularity, the exoskeleton can be used as a full lower-limb orthosis, a single-joint orthosis in any of the three joints, and a two-joint orthosis in a combination of any of the two joints. By employing a simple torque control strategy, the exoskeleton can be used to deliver user-specific assistance, both in gait rehabilitation and in assisting people suffering musculoskeletal impairments. The result of the presented BioMot efforts is a low-footprint exoskeleton with powerful compliant actuators, simple, yet effective torque controller and easily adjustable flexible structure.

  1. The joint flanker effect and the joint Simon effect: On the comparability of processes underlying joint compatibility effects.

    PubMed

    Dittrich, Kerstin; Bossert, Marie-Luise; Rothe-Wulf, Annelie; Klauer, Karl Christoph

    2017-09-01

    Previous studies observed compatibility effects in different interference paradigms such as the Simon and flanker task even when the task was distributed across two co-actors. In both Simon and flanker tasks, performance is improved in compatible trials relative to incompatible trials if one actor works on the task alone as well as if two co-actors share the task. These findings have been taken to indicate that actors automatically co-represent their co-actor's task. However, recent research on the joint Simon and joint flanker effect suggests alternative non-social interpretations. To which degree both joint effects are driven by the same underlying processes is the question of the present study, and it was scrutinized by manipulating the visibility of the co-actor. While the joint Simon effect was not affected by the visibility of the co-actor, the joint flanker effect was reduced when participants did not see their co-actors but knew where the co-actors were seated. These findings provide further evidence for a spatial interpretation of the joint Simon effect. In contrast to recent claims, however, we propose a new explanation of the joint flanker effect that attributes the effect to an impairment in the focusing of spatial attention contingent on the visibility of the co-actor.

  2. The prevalence of sacroiliac joint degeneration in asymptomatic adults.

    PubMed

    Eno, Jonathan-James T; Boone, Christopher R; Bellino, Michael J; Bishop, Julius A

    2015-06-03

    Degenerative changes of the sacroiliac joint have been implicated as a cause of lower back pain in adults. The purpose of this study was to determine the prevalence of sacroiliac joint degeneration in asymptomatic patients. Five hundred consecutive pelvic computed tomography (CT) scans, made at a tertiary-care medical center, of patients with no history of pain in the lower back or pelvic girdle were retrospectively reviewed and analyzed for degenerative changes of the sacroiliac joint. After exclusion criteria were applied, 373 CT scans (746 sacroiliac joints) were evaluated for degenerative changes. Regression analysis was used to determine the association between age and the degree of sacroiliac joint degeneration. The prevalence of sacroiliac joint degeneration was 65.1%, with substantial degeneration occurring in 30.5% of asymptomatic subjects. The prevalence steadily increased with age, with 91% of subjects in the ninth decade of life displaying degenerative changes. Radiographic evidence of sacroiliac joint degeneration is highly prevalent in the asymptomatic population and is associated with age. Caution must be exercised when attributing lower back or pelvic girdle pain to sacroiliac joint degeneration seen on imaging. Prognostic Level IV. See Instructions for Authors for a complete description of levels of evidence. Copyright © 2015 by The Journal of Bone and Joint Surgery, Incorporated.

  3. Microstructure and mechanical properties of China low activation martensitic steel joint by TIG multi-pass welding with a new filler wire

    NASA Astrophysics Data System (ADS)

    Huang, Bo; Zhang, Junyu; Wu, Qingsheng

    2017-07-01

    Tungsten Inner Gas (TIG) welding is employed for joining of China low activation martensitic (CLAM) steel. A new filler wire was proposed, and the investigation on welding with various heat input and welding passes were conducted to lower the tendency towards the residual of δ ferrite in the joint. With the optimized welding parameters, a butt joint by multi-pass welding with the new filler wire was prepared to investigate the microstructure and mechanical properties. The microstructure of the joint was observed by optical microscope (OM) and scanning electron microscope (SEM). The hardness, Charpy impact and tensile tests of the joint were implemented at room temperature (25 °C). The results revealed that almost full martensite free from ferrite in the joints were obtained by multipass welding with the heat input of 2.26 kJ/mm. A certain degree of softening occurred at the heat affected zone of the joint according to the results of tensile and hardness tests. The as welded joints showed brittle fracture in the impact tests. However, the joints showed toughness fracture after tempering and relatively better comprehensive performance were achieved when the joints were tempered at 740 °C for 2 h.

  4. Development and validation of a computational model of the knee joint for the evaluation of surgical treatments for osteoarthritis

    PubMed Central

    Mootanah, R.; Imhauser, C.W.; Reisse, F.; Carpanen, D.; Walker, R.W.; Koff, M.F.; Lenhoff, M.W.; Rozbruch, S.R.; Fragomen, A.T.; Dewan, Z.; Kirane, Y.M.; Cheah, Pamela A.; Dowell, J.K.; Hillstrom, H.J.

    2014-01-01

    A three-dimensional (3D) knee joint computational model was developed and validated to predict knee joint contact forces and pressures for different degrees of malalignment. A 3D computational knee model was created from high-resolution radiological images to emulate passive sagittal rotation (full-extension to 65°-flexion) and weight acceptance. A cadaveric knee mounted on a six-degree-of-freedom robot was subjected to matching boundary and loading conditions. A ligament-tuning process minimised kinematic differences between the robotically loaded cadaver specimen and the finite element (FE) model. The model was validated by measured intra-articular force and pressure measurements. Percent full scale error between EE-predicted and in vitro-measured values in the medial and lateral compartments were 6.67% and 5.94%, respectively, for normalised peak pressure values, and 7.56% and 4.48%, respectively, for normalised force values. The knee model can accurately predict normalised intra-articular pressure and forces for different loading conditions and could be further developed for subject-specific surgical planning. PMID:24786914

  5. Development and validation of a computational model of the knee joint for the evaluation of surgical treatments for osteoarthritis.

    PubMed

    Mootanah, R; Imhauser, C W; Reisse, F; Carpanen, D; Walker, R W; Koff, M F; Lenhoff, M W; Rozbruch, S R; Fragomen, A T; Dewan, Z; Kirane, Y M; Cheah, K; Dowell, J K; Hillstrom, H J

    2014-01-01

    A three-dimensional (3D) knee joint computational model was developed and validated to predict knee joint contact forces and pressures for different degrees of malalignment. A 3D computational knee model was created from high-resolution radiological images to emulate passive sagittal rotation (full-extension to 65°-flexion) and weight acceptance. A cadaveric knee mounted on a six-degree-of-freedom robot was subjected to matching boundary and loading conditions. A ligament-tuning process minimised kinematic differences between the robotically loaded cadaver specimen and the finite element (FE) model. The model was validated by measured intra-articular force and pressure measurements. Percent full scale error between FE-predicted and in vitro-measured values in the medial and lateral compartments were 6.67% and 5.94%, respectively, for normalised peak pressure values, and 7.56% and 4.48%, respectively, for normalised force values. The knee model can accurately predict normalised intra-articular pressure and forces for different loading conditions and could be further developed for subject-specific surgical planning.

  6. Development of a revolute-joint robot for the precision positioning of an x-ray detector

    NASA Astrophysics Data System (ADS)

    Preissner, Curt A.; Royston, Thomas J.; Shu, Deming

    2003-10-01

    This paper profiles the initial phase in the development of a six degree-of-freedom robot, with 1 μm dynamic positioning uncertainty, for the manipulation of x-ray detectors or test specimens at the Advanced Photon Source (APS). While revolute-joint robot manipulators exhibit a smaller footprint along with increased positioning flexibility compared to Cartesian manipulators, commercially available revolute-joint manipulators do not meet our size, positioning, or environmental specifications. Currently, a robot with 20 μm dynamic positioning uncertainty is functioning at the APS for cryogenic crystallography sample pick-and-place operation. Theoretical, computational and experimental procedures are being used to (1) identify and (2) simulate the dynamics of the present robot system using a multibody approach, including the mechanics and control architecture, and eventually to (3) design an improved version with a 1 μm dynamic positioning uncertainty. We expect that the preceding experimental and theoretical techniques will be useful design and analysis tools as multi-degree-of-freedom manipulators become more prevalent on synchrotron beamlines.

  7. Effects of a retention intervention program for associate degree nursing students.

    PubMed

    Fontaine, Karen

    2014-01-01

    To evaluate the effects of a retention intervention program on nursing students' persistence in obtaining an associate's degree. An associate degree nursing program at a large community college used a three-year grant from the US Department of Labor to create a program to improve retention of nursing students. Seven retention interventions (stipends, learning communities, comprehensive orientation, individualized academic planning, counseling, peer tutoring, and community nurse mentoring) were provided to participants. Correlational analyses were conducted between demographic variables and degree completion and between individual intervention program participation and degree completion. The program produced a statistically significant improvement in retention, but no specific intervention or mixture of interventions was significantly correlated with retention. Retention programs must be comprehensive, integrated efforts in order to increase the degree completion rate.

  8. MD/MBA Students: An Analysis of Medical Student Career Choice.

    PubMed

    Sherrill, Windsor Westbrook

    2004-12-01

    An increasing number of medical schools are offering dual degree MD/MBA programs. Career choices and factors influencing students to enter these programs provide an indicator of the roles in which dual degree students will serve in health care as well as the future of dual degree programs. Using career choice theory as a conceptual framework, career goals and factors influencing decisions to enter dual degree programs were assessed among dual degree medical students. Students enrolled at dual degree programs at six medical schools were surveyed and interviewed. A control group of traditional medical students was also surveyed. Factors influencing students to seek both medical and business training are varied but are often related to a desire for leadership opportunities, concerns about change in medicine and job security and personal career goals. Most students expect to combine clinical and administrative roles. Students entering these programs do so for a variety of reasons and plan diverse careers. These findings can provide guidance for program development and recruitment for dual degree medical education programs.

  9. Seal Joint Analysis and Design for the Ares-I Upper Stage LOX Tank

    NASA Technical Reports Server (NTRS)

    Phillips, Dawn R.; Wingate, Robert J.

    2011-01-01

    The sealing capability of the Ares-I Upper Stage liquid oxygen tank-to-sump joint is assessed by analyzing the deflections of the joint components. Analyses are performed using three-dimensional symmetric wedge finite element models and the ABAQUS commercial finite element software. For the pressure loads and feedline interface loads, the analyses employ a mixed factor of safety approach to comply with the Constellation Program factor of safety requirements. Naflex pressure-assisted seals are considered first because they have been used successfully in similar seal joints in the Space Shuttle External Tank. For the baseline sump seal joint configuration with a Naflex seal, the predicted joint opening greatly exceeds the seal design specification. Three redesign options of the joint that maintain the use of a Naflex seal are studied. The joint openings for the redesigned seal joints show improvement over the baseline configuration; however, these joint openings still exceed the seal design specification. RACO pressure-assisted seals are considered next because they are known to also be used on the Space Shuttle External Tank, and the joint opening allowable is much larger than the specification for the Naflex seals. The finite element models for the RACO seal analyses are created by modifying the models that were used for the Naflex seal analyses. The analyses show that the RACO seal may provide sufficient sealing capability for the sump seal joint. The results provide reasonable data to recommend the design change and plan a testing program to determine the capability of RACO seals in the Ares-I Upper Stage liquid oxygen tank sump seal joint.

  10. Four Weeks of Balance Training does not Affect Ankle Joint Stiffness in Subjects with Unilateral Chronic Ankle Instability

    PubMed Central

    Jain, Tarang Kumar; Wauneka, Clayton N.; Liu, Wen

    2016-01-01

    Background Balance training has been shown to be effective in preventing ankle sprain recurrences in subjects with chronic ankle instability (CAI) but the biomechanical pathways underlying the clinical outcomes are still unknown. This study was conducted to determine if a 4-week balance training intervention can alter the mechanical characteristics in ankles with CAI. Methods Twenty-two recreationally active subjects with unilateral CAI were randomized to either a control (n = 11, 35.1 ± 9.3 years) or intervention (n = 11, 33.5 ± 6.6 years) group. Subjects in the intervention group were trained on the affected limb with static and dynamic components using a Biodex balance stability system for 4-weeks. The ankle joint stiffness and neutral zone in inversion and eversion directions on the involved and uninvolved limbs was measured at baseline and post-intervention using a dynamometer. Results At baseline, the mean values of the inversion stiffness (0.69 ± 0.37 Nm/degree) in the involved ankle was significantly lower (p < 0.011, 95% CI [0.563, 0.544]) than that of uninvolved contralateral ankle (0.99 ± 0.41 Nm/degree). With the available sample size, the eversion stiffness, inversion neutral zone, and eversion neutral zone were not found to be significantly different between the involved and uninvolved contralateral ankles. The 4-week balance training intervention failed to show any significant effect on the passive ankle stiffness and neutral zones in inversion and eversion. Conclusion Decreased inversion stiffness in the involved chronic unstable ankle was found that of uninvolved contralateral ankle. The 4-week balance training program intervention was ineffective in altering the mechanical characteristics of ankles with CAI. Level of evidence Randomized controlled clinical trial; Level of evidence, 1. PMID:27642647

  11. The Joint Winter Runway Friction Measurement Program: NASA Perspective

    NASA Technical Reports Server (NTRS)

    Yager, Thomas J.

    1996-01-01

    Some background information is given together with the scope and objectives of the 5-year, Joint National Aeronautics & Space Administration (NASA)/Transport Canada (TC)/Federal Aviation Administration (FAA) Winter Runway Friction Measurement Program. The range of the test equipment, the selected test sites and a tentative test program schedule are described. NASA considers the success of this program critical in terms of insuring adequate ground handling performance capability in adverse weather conditions for future aircraft being designed and developed as well as improving the safety of current aircraft ground operations.

  12. Second-Degree Learners in Associate Degree Nursing Programs: Characteristics and Progression Success

    ERIC Educational Resources Information Center

    McGinley, Patricia M.

    2013-01-01

    Second degree learners are attending associate (ADN), accelerated baccalaureate (BSN), and entry level masters (ELM) degree nursing programs. There is limited data related to the socio-demographic characteristics and graduation success rates of students attending accelerated BSN or ELM programs and no data related to second-degree learners…

  13. Two Universities, Two Degrees: A Dual Degree Program for Pharmacists.

    ERIC Educational Resources Information Center

    Milio, Frank

    2001-01-01

    Describes a dual degree program between Towson University and the University of Maryland School of Pharmacy, which allows a student to receive both a B.S. degree in Medicinal Chemistry and a Doctor of Pharmacy degree in a combined 7-year program. It also allows flexibility in pursuing alternate career goals. (EV)

  14. A Study of H-Reflexes in Subjects with Acute Ankle Inversion Injuries

    DTIC Science & Technology

    1996-12-09

    stress to the injured ankle at heel- strike .(57) Any increased inversion stress by way of joint loading in the presence of compromised joint...the present study, may play a role in decreasing the degree of calcaneal inversion just prior to heel- strike and minimize the stress on the lateral...Presentation: * Significant edema/ecchymosis on lateral and medial aspects of ankle. * Possible pitting edema on forefoot (several days post- injury

  15. Kinematic control of redundant robots and the motion optimizability measure.

    PubMed

    Li, L; Gruver, W A; Zhang, Q; Yang, Z

    2001-01-01

    This paper treats the kinematic control of manipulators with redundant degrees of freedom. We derive an analytical solution for the inverse kinematics that provides a means for accommodating joint velocity constraints in real time. We define the motion optimizability measure and use it to develop an efficient method for the optimization of joint trajectories subject to multiple criteria. An implementation of the method for a 7-dof experimental redundant robot is present.

  16. An Analysis of Operational Availability of Brazilian Navy and Argentine Air Force A-4 Fleets Using Simulation Modeling

    DTIC Science & Technology

    1999-12-01

    Escuela de Ingenieria Aeronäutica, 1982 Submitted in partial fulfillment of the requirements for the degree of MASTER OF SCD2NCE IN MANAGEMENT...defense industry , including joint production, technical data sharing, and reciprocal subcontracting in the joint development and production of a...34Foundations of Manufacturing Management". 15. Barton, Russell R., Simulation Metamodels, Department of Industrial and Manufacturing Engineering, The

  17. The World According to de Finetti: On de Finetti's Theory of Probability and Its Application to Quantum Mechanics

    NASA Astrophysics Data System (ADS)

    Berkovitz, Joseph

    Bruno de Finetti is one of the founding fathers of the subjectivist school of probability, where probabilities are interpreted as rational degrees of belief. His work on the relation between the theorems of probability and rationality is among the corner stones of modern subjective probability theory. De Finetti maintained that rationality requires that degrees of belief be coherent, and he argued that the whole of probability theory could be derived from these coherence conditions. De Finetti's interpretation of probability has been highly influential in science. This paper focuses on the application of this interpretation to quantum mechanics. We argue that de Finetti held that the coherence conditions of degrees of belief in events depend on their verifiability. Accordingly, the standard coherence conditions of degrees of belief that are familiar from the literature on subjective probability only apply to degrees of belief in events which could (in principle) be jointly verified; and the coherence conditions of degrees of belief in events that cannot be jointly verified are weaker. While the most obvious explanation of de Finetti's verificationism is the influence of positivism, we argue that it could be motivated by the radical subjectivist and instrumental nature of probability in his interpretation; for as it turns out, in this interpretation it is difficult to make sense of the idea of coherent degrees of belief in, and accordingly probabilities of unverifiable events. We then consider the application of this interpretation to quantum mechanics, concentrating on the Einstein-Podolsky-Rosen experiment and Bell's theorem.

  18. Race-Neutral Campuses in Urban Areas: A Follow-Up Report on Merger and Joint Planning.

    ERIC Educational Resources Information Center

    Godard, James M.

    Developments during the 1980-81 academic year concerning mergers/joint planning projects designed to eliminate unnecessary duplication of programs between historically black and white colleges in Nashville, Tennessee; Savannah, Georgia; and Norfolk, Virginia are examined. Site visits were made during the second year of operation of the programs at…

  19. A joint labor-management hazard communication training program: a case study in worker health and safety training.

    PubMed

    Robins, T G; Hugentobler, M K; Kaminski, M; Klitzman, S

    1994-01-01

    The 1983 OSHA Hazard Communication Standard requires training of employees exposed to hazardous chemicals. The authors provide a detailed look at the successes and failures of a joint labor-management training program that was designed to bring a firm with more than 50 manufacturing facilities into compliance with the standard.

  20. Durability assessments of concrete using electrical properties and acoustic emission testing

    NASA Astrophysics Data System (ADS)

    Todak, Heather N.

    Premature damage deterioration has been observed in pavement joints throughout the Midwestern region of the United States. Over time, severe joint damage creates a transportation safety concern and the necessary repairs can be an extreme economic burden. The deterioration is due in part to freeze-thaw damage associated with fluid accumulation at the pavement joints. This very preventable problem is an indication that current specifications and construction practices for freeze-thaw durability of concrete are inadequate. This thesis serves to create a better understanding of moisture ingress, freeze-thaw damage mechanisms, and the effect of variations in mixture properties on freeze-thaw behavior of concrete. The concepts of the nick point degree of saturation, sorptivity rates, and critical degree of saturation are discussed. These factors contribute to service life, defined in this study as the duration of time a concrete element remains below levels of critical saturation which are required for damage development to initiate. A theoretical model and a simple experimental procedure are introduced which help determine the nick point for a series of 32 concrete mixtures with unique mixture proportions and air entrainment properties. This simple experimental procedure is also presented as a method to measure important electrical properties in order to establish the formation factor, a valuable measure of concrete transport properties. The results of freeze-thaw testing with acoustic emission monitoring are presented to help understand and quantify damage development in concrete specimens when conditioned to various degrees of saturation. This procedure was used to study the relationship between air entrainment properties and the critical degree of saturation. Applying the concepts of degree of saturation and sorptivity, a performance-based model is proposed as a new approach to specifications for freeze-thaw durability. Finally, a conceptual model is presented to illustrate the effect of various changes in mixture proportions and air void properties on service life.

  1. Effect of Radiofrequency Denervation on Pain Intensity Among Patients With Chronic Low Back Pain: The Mint Randomized Clinical Trials.

    PubMed

    Juch, Johan N S; Maas, Esther T; Ostelo, Raymond W J G; Groeneweg, J George; Kallewaard, Jan-Willem; Koes, Bart W; Verhagen, Arianne P; van Dongen, Johanna M; Huygen, Frank J P M; van Tulder, Maurits W

    2017-07-04

    Radiofrequency denervation is a commonly used treatment for chronic low back pain, but high-quality evidence for its effectiveness is lacking. To evaluate the effectiveness of radiofrequency denervation added to a standardized exercise program for patients with chronic low back pain. Three pragmatic multicenter, nonblinded randomized clinical trials on the effectiveness of minimal interventional treatments for participants with chronic low back pain (Mint study) were conducted in 16 multidisciplinary pain clinics in the Netherlands. Eligible participants were included between January 1, 2013, and October 24, 2014, and had chronic low back pain, a positive diagnostic block at the facet joints (facet joint trial, 251 participants), sacroiliac joints (sacroiliac joint trial, 228 participants), or a combination of facet joints, sacroiliac joints, or intervertebral disks (combination trial, 202 participants) and were unresponsive to conservative care. All participants received a 3-month standardized exercise program and psychological support if needed. Participants in the intervention group received radiofrequency denervation as well. This is usually a 1-time procedure, but the maximum number of treatments in the trial was 3. The primary outcome was pain intensity (numeric rating scale, 0-10; whereby 0 indicated no pain and 10 indicated worst pain imaginable) measured 3 months after the intervention. The prespecified minimal clinically important difference was defined as 2 points or more. Final follow-up was at 12 months, ending October 2015. Among 681 participants who were randomized (mean age, 52.2 years; 421 women [61.8%], mean baseline pain intensity, 7.1), 599 (88%) completed the 3-month follow-up, and 521 (77%) completed the 12-month follow-up. The mean difference in pain intensity between the radiofrequency denervation and control groups at 3 months was -0.18 (95% CI, -0.76 to 0.40) in the facet joint trial; -0.71 (95% CI, -1.35 to -0.06) in the sacroiliac joint trial; and -0.99 (95% CI, -1.73 to -0.25) in the combination trial. In 3 randomized clinical trials of participants with chronic low back pain originating in the facet joints, sacroiliac joints, or a combination of facet joints, sacroiliac joints, or intervertebral disks, radiofrequency denervation combined with a standardized exercise program resulted in either no improvement or no clinically important improvement in chronic low back pain compared with a standardized exercise program alone. The findings do not support the use of radiofrequency denervation to treat chronic low back pain from these sources. trialregister.nl Identifier: NTR3531.

  2. Brief 74 Nuclear Engineering Enrollments and Degrees Survey, 2014 Data

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    None, None

    2015-03-15

    The 2014 survey includes degrees granted between September 1, 2013 and August 31, 2014, and enrollments for fall 2014. There are three academic programs new to this year's survey. Thirty-five academic programs reported having nuclear engineering programs during 2014, and data were provided by all thirty-five. The enrollments and degrees data include students majoring in nuclear engineering or in an option program equivalent to a major. Two nuclear engineering programs have indicated that health physics option enrollments and degrees are also reported in the health physics enrollments and degrees survey.

  3. [Acetabuloplasty in cerebral palsy patients].

    PubMed

    Schejbalová, A; Chládek, P

    2007-12-01

    Acetabuloplasty as an isolated surgical procedure is one of the options allowing for hip joint realignment in cerebral palsy patients. Also, it is often involved in combined techniques used for hip joint reconstruction. In the years 2004-2005, 20 cerebral palsy patients, aged from 4 to 13 years, were indicated for acetabuloplasty. The group included children with varying degrees of locomotor disability, ranging from inability even to crawl to ability to move without support (stages 2 to 7 on the Vojta scale). Clinical and X-ray findings were evaluated at 2 and 6 months after surgical treatment and then every 6 months of follow- up. The X-ray finding was described by means of CE angle (Wiberg) values and hip migration percentage. Of 21 hips (in 20 patients) treated by acetabuloplasty in our modification during the two years, one developed recurrent dorsal dislocation which was possible to deal with by conservative treatment. All hips pre-operatively categorized as group B or group C were post-operatively assessed as group A hips. Subsequent migration of the acetabulum occurred within a year in two patients with distinct original asymmetry. Surgery on bones comprising the hip joint in cerebral palsy patients is indicated according to the degree of lateral migration and changes in hip joint geometry involving the proximal femur or pelvis, or in combination with open reduction. Acetabuloplasty is indicated as an isolated procedure or as part of combined techniques leading to a better coverage of the hip joint. It is recommended for children up to 10 years of age, but also older ones, in whom the flexibility of child bones enables us to do without internal osteosynthesis. Acetabuloplasty is indicated as an isolated surgical procedure for hip joint subluxation in cerebral palsy children. It has no adverse effects on hip abductors that, in cerebral palsy patients, are 85 % insufficient. It improves hip joint symmetry and helps to avoid more demanding reconstructive or palliative procedures. Key words: cerebral palsy, acetabuloplasty, lateral migration.

  4. Tennis players show a lower coactivation of the elbow antagonist muscles during isokinetic exercises.

    PubMed

    Bazzucchi, Ilenia; Riccio, Maria Elena; Felici, Francesco

    2008-10-01

    Previous studies have suggested that muscle coactivation could be reduced by a recurrent activity (training, daily activities). If this was correct, skilled athletes should show a specific muscle activation pattern with a low level of coactivation of muscles which are typically involved in their discipline. In particular, the aim of this study was to verify the hypothesis that the amount of antagonist activation of biceps brachii (BB) and triceps brachii (TB) is different between tennis players and non-players individuals during maximal isokinetic contractions. Ten young healthy men and eight male tennis players participated in the study. The surface electromyographic signals (sEMG) were recorded from the BB and TB muscles during three maximal voluntary isometric contractions (MVC) of elbow flexors and extensors and a set of three maximal elbow flexions and extensions at 15 degrees , 30 degrees , 60 degrees , 120 degrees , 180 degrees and 240 degrees /s. Normalized root mean square (RMS) of sEMG was calculated as an index of sEMG amplitude. Antagonist activation (%RMSmax) of TB was significantly lower in tennis players (from 14.0+/-7.9% at MVC to 16.3+/-8.9% at 240 degrees /s) with respect to non-players (from 27.7+/-19.7% at MVC to 38.7+/-17.6% at 240 degrees /s) at all angular velocities. Contrary to non-players, tennis players did not show any difference in antagonist activation between BB and TB muscles. Tennis players, with a constant practice in controlling forces around the elbow joint, learn how to reduce coactivation of muscles involved in the control of this joint. This has been shown by the lower antagonist muscular activity of triceps brachii muscle during isokinetic elbow flexion found in tennis players with respect to non-players.

  5. [Demographic features and difficulties in rehabilitation in patients referred to hand rehabilitation unit for phalangeal fractures].

    PubMed

    Sahin, Füsun; Yücel, Serap Dalgiç; Yilmaz, Figen; Ergöz, Ernur; Kuran, Banu

    2006-01-01

    We evaluated demographic and occupational features of patients with phalangeal fractures of the hand, etiologies and types of injuries, and the results of rehabilitation. The study included 91 fingers of 62 patients (54 males, 8 females; mean age 28+/-13 years; range 4 to 59 years) who were referred to our hand rehabilitation unit for phalangeal fractures. Demographic features, the cause and localization of injury, the type of surgery, time from surgery to rehabilitation, and the follow-up period were determined. At the end of rehabilitation, range of motion (ROM) of the phalangeal joint and total ROM of the injured fingers were assessed using the Strickland-Glogovac rating system. A great majority of injuries were caused by work accidents, followed by sport injuries and falls occurring in students. Sixty patients (96.8%) were right-handed. The fractures occurred in the dominant hand in 29 patients (46.8%). The majority of patients (n=45) were primary school graduates. The most common mechanism of injury was accidents related to heavy work machinery (n=18). The most commonly injured finger and the phalanx were the third finger (n=25, 27.5%) and the proximal phalanx (n=59, 56.7%), respectively. Only 27 patients (43.6%) had a sufficient follow-up with a mean of 79.7+/-46.6 days (range 30 to 254 days). Following rehabilitation, the mean ROM and the total ROM were 45.0+/-22.9 degrees and 63.3+/-16.1 degrees for the injured joint and the thumb, and 31.3+/-22.5 degrees and 122+/-60.3 degrees for the injured joint and the other fingers, respectively. Our data provide important insight into appropriate treatment and rehabilitation of phalangeal fractures, in particular, shortcomings in the treatment and follow-up.

  6. Microstructural and mechanical characterization of CO{sub 2} laser and gas tungsten arc welds of an Al-Li-Cu alloy 2195

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hou, K.H.; Baeslack, W.A. III; Szabo, A.

    1994-12-31

    Lithium-containing aluminum alloys offer an attractive combination of low density and high strength and stiffness and have been the focus of vigorous research for their promising aerospace applications. To achieve the full potential advantages in using these alloys, the integrity of welded joints, both n the fusion zone and the heat-affected zone, must be ensured. In the present study, Weldalite{sup TM} 049 (designated as alloy 2195) with nominal composition of Al-1.0Li-4.0Cu-0.4Mg0.4Ag-0.14Zr (wt%) was welded autogenously using the gas tungsten-arc (GTA) and CO{sub 2} laser beam (LB) welding processes. The average ultimate tensile strengths for as-welded, 160{degrees}C/16 h-aged, and 190{degrees}C/16 h-agedmore » GTA welds were 296.4 MPa, 304.6 MPa, and 336.8 MPa, and corresponded to joint efficiencies of 61.4%, 48.1% and 56.0%, respectively. Porosity was found occasionally in the laser welds and slightly affected the performance of the aluminum weldments. For laser welds, average ultimate tensile strengths and corresponding joint efficiencies for a-welded, 160{degrees}C/16 h-aged, and 190{degrees}C/16 h-aged weldments were 293.2 MPa (60.8%) 305.9 MPa (48.3%), and 331.0 MPa (55.0%), respectively. Scanning electron fractography revealed that failure of the GTA and LB tensile specimens occurred either within the weld metal or along the fusion boundary. The latter was related to the existence of an equiaxed band along the fusion boundary.« less

  7. Analytical formulation of orbiter-payload models coupled by trunnion joints with Coulomb friction

    NASA Technical Reports Server (NTRS)

    Liu, Frank C.

    1987-01-01

    An orbiter and its payload substructure are linked together by five trunnion joints which have thirty degrees-of-freedom. Geometric compatibility conditions require fourteen of the interface physical coordinates of the orbiter and payload to be equal to each other and the remaining sixteen are free to have relative motions under Coulomb friction. The component modes synthesis method using fourteen inertia relief attachment modes for the formulation of the coupled system is presented. The exact nonlinear friction function is derived based on the characteristics of the joints. Formulation is applicable to an orbiter that carries any number of payload substructures.

  8. Analytical formulation of orbiter-payload coupled by trunnion joints with Coulomb friction

    NASA Technical Reports Server (NTRS)

    Liu, Frank C.

    1986-01-01

    An orbiter and its payload substructure are linked together by five trunnion joints which have thirty degrees-of-freedom. Geometric compatibility conditions require fourteen of the interface physical coordinates of the orbiter and payload to be equal to each other and the remaining sixteen are free to have relative motions under Coulomb friction. The component modes synthesis method using fourteen inertia relief attachment modes for the formulation of the coupled system is presented. The exact nonlinear friction function is derived based on the characteristics of the joints. Formulation is applicable to an orbiter that carries any number of payload substructures.

  9. Design/analysis of the JWST ISIM bonded joints for survivability at cryogenic temperatures

    NASA Astrophysics Data System (ADS)

    Bartoszyk, Andrew; Johnston, John; Kaprielian, Charles; Kuhn, Jonathan; Kunt, Cengiz; Rodini, Benjamin; Young, Daniel

    2005-08-01

    A major design and analysis challenge for the JWST ISIM structure is thermal survivability of metal/composite adhesively bonded joints at the cryogenic temperature of 30K (-405°F). Current bonded joint concepts include internal invar plug fittings, external saddle titanium/invar fittings and composite gusset/clip joints all bonded to hybrid composite tubes (75mm square) made with M55J/954-6 and T300/954-6 prepregs. Analytical experience and design work done on metal/composite bonded joints at temperatures below that of liquid nitrogen are limited and important analysis tools, material properties, and failure criteria for composites at cryogenic temperatures are sparse in the literature. Increasing this challenge is the difficulty in testing for these required tools and properties at cryogenic temperatures. To gain confidence in analyzing and designing the ISIM joints, a comprehensive joint development test program has been planned and is currently running. The test program is designed to produce required analytical tools and develop a composite failure criterion for bonded joint strengths at cryogenic temperatures. Finite element analysis is used to design simple test coupons that simulate anticipated stress states in the flight joints; subsequently, the test results are used to correlate the analysis technique for the final design of the bonded joints. In this work, we present an overview of the analysis and test methodology, current results, and working joint designs based on developed techniques and properties.

  10. Aircrew Sizing Survey 2011

    DTIC Science & Technology

    2014-10-01

    Resource JSF Joint Strike Fighter JPATS Joint Primary Aircraft Training System USMC United States Marine Corps USAF United States Air Force LIST...Surface Anthropometry Resource (CAESAR) was developed for the Joint Strike Fighter (JSF) program. The ACSS was intended to replace the JSF-CAESAR...an aircrew sample was made in 2003 by Hudson et al. They extracted a subset, named JSF CAESAR (Joint Strike Fighter), from the Civilian American and

  11. 45 CFR 2400.41 - Degree programs.

    Code of Federal Regulations, 2013 CFR

    2013-10-01

    ... degree in education that permits a concentration in American history, American government, social studies, or political science. Graduate degrees under which study is excluded from fellowship support are... FELLOWSHIP PROGRAM REQUIREMENTS Graduate Study § 2400.41 Degree programs. (a) Fellows may pursue a master's...

  12. 45 CFR 2400.41 - Degree programs.

    Code of Federal Regulations, 2012 CFR

    2012-10-01

    ... degree in education that permits a concentration in American history, American government, social studies, or political science. Graduate degrees under which study is excluded from fellowship support are... FELLOWSHIP PROGRAM REQUIREMENTS Graduate Study § 2400.41 Degree programs. (a) Fellows may pursue a master's...

  13. 45 CFR 2400.41 - Degree programs.

    Code of Federal Regulations, 2010 CFR

    2010-10-01

    ... degree in education that permits a concentration in American history, American government, social studies, or political science. Graduate degrees under which study is excluded from fellowship support are... FELLOWSHIP PROGRAM REQUIREMENTS Graduate Study § 2400.41 Degree programs. (a) Fellows may pursue a master's...

  14. 45 CFR 2400.41 - Degree programs.

    Code of Federal Regulations, 2011 CFR

    2011-10-01

    ... degree in education that permits a concentration in American history, American government, social studies, or political science. Graduate degrees under which study is excluded from fellowship support are... FELLOWSHIP PROGRAM REQUIREMENTS Graduate Study § 2400.41 Degree programs. (a) Fellows may pursue a master's...

  15. 45 CFR 2400.41 - Degree programs.

    Code of Federal Regulations, 2014 CFR

    2014-10-01

    ... degree in education that permits a concentration in American history, American government, social studies, or political science. Graduate degrees under which study is excluded from fellowship support are... FELLOWSHIP PROGRAM REQUIREMENTS Graduate Study § 2400.41 Degree programs. (a) Fellows may pursue a master's...

  16. Low-Friction, Low-Profile, High-Moment Two-Axis Joint

    NASA Technical Reports Server (NTRS)

    Lewis, James L.; Le, Thang; Carroll, Monty B.

    2010-01-01

    The two-axis joint is a mechanical device that provides two-degrees-of-freedom motion between connected components. A compact, moment-resistant, two-axis joint is used to connect an electromechanical actuator to its driven structural members. Due to the requirements of the overall mechanism, the joint has a low profile to fit within the allowable space, low friction, and high moment-reacting capability. The mechanical arrangement of this joint can withstand high moments when loads are applied. These features allow the joint to be used in tight spaces where a high load capability is required, as well as in applications where penetrating the mounting surface is not an option or where surface mounting is required. The joint consists of one base, one clevis, one cap, two needle bearings, and a circular shim. The base of the joint is the housing (the base and the cap together), and is connected to the grounding structure via fasteners and a bolt pattern. Captive within the housing, between the base and the cap, are the rotating clevis and the needle bearings. The clevis is attached to the mechanical system (linear actuator) via a pin. This pin, and the rotational movement of the clevis with respect to the housing, provides two rotational degrees of freedom. The larger diameter flange of the clevis is sandwiched between a pair of needle bearings, one on each side of the flange. During the assembly of the two-axis joint, the circular shims are used to adjust the amount of preload that is applied to the needle bearings. The above arrangement enables the joint to handle high moments with minimal friction. To achieve the high-moment capability within a low-profile joint, the use of depth of engagement (like that of a conventional rotating shaft) to react moment is replaced with planar engagement parallel to the mounting surface. The needle bearings with the clevis flange provide the surface area to react the clevis loads/moments into the joint housing while providing minimal friction during rotation. The diameter of the flange and the bearings can be increased to react higher loads and still maintain a compact surface mounting capability. This type of joint can be used in a wide variety of mechanisms and mechanical systems. It is especially effective where precise, smooth, continuous motion is required. For example, the joint can be used at the end of a linear actuator that is required to extend and rotate simultaneously. The current design application is for use in a spacecraft docking-system capture mechanism. Other applications might include industrial robotic or assembly line apparatuses, positioning systems, or in the motion-based simulator industry that employs complex, multi-axis manipulators for various types of motions.

  17. Report on an Investigation into an Entry Level Clinical Doctorate for the Genetic Counseling Profession and a Survey of the Association of Genetic Counseling Program Directors.

    PubMed

    Reiser, Catherine; LeRoy, Bonnie; Grubs, Robin; Walton, Carol

    2015-10-01

    The master's degree is the required entry-level degree for the genetic counseling profession in the US and Canada. In 2012 the Association of Genetic Counseling Program Directors (AGCPD) passed resolutions supporting retention of the master's as the entry-level and terminal degree and opposing introduction of an entry-level clinical doctorate (CD) degree. An AGCPD workgroup surveyed directors of all 34 accredited training programs with the objective of providing the Genetic Counseling Advanced Degrees Task Force (GCADTF) with information regarding potential challenges if master's programs were required to transition to an entry-level CD. Program demographics, projected ability to transition to an entry-level CD, factors influencing ability to transition, and potential effects of transition on programs, students and the genetic counseling workforce were characterized. Two programs would definitely be able to transition, four programs would close, thirteen programs would be at risk to close and fourteen programs would probably be able to transition with varying degrees of difficulty. The most frequently cited limiting factors were economic, stress on clinical sites, and administrative approval of a new degree/program. Student enrollment under an entry-level CD model was projected to decrease by 26.2 %, negatively impacting the workforce pipeline. The results further illuminate and justify AGCPD's position to maintain the master's as the entry-level degree.

  18. Recognition of Academic Qualifications in Transnational Higher Education and Challenges for Recognizing a Joint Degree in Europe and Asia

    ERIC Educational Resources Information Center

    Hou, Angela Yung-Chi; Morse, Robert; Wang, Wayne

    2017-01-01

    Since the 1950s, the Council of Europe has established conventions and information networks to enhance student mobility and qualification recognition in Europe. In contrast, Asia did not take this issue into consideration until 1983, when the UNESCO Regional Convention on the Recognition of Studies, Diplomas and Degrees in Higher Education in Asia…

  19. [Dual x-ray absorptiometry assessment of bone density of the proximal tibia in advanced-stage degenerative disease of the knee].

    PubMed

    Hulet, C; Sabatier, J P; Schiltz, D; Locker, B; Marcelli, C; Vielpeau, C

    2001-02-01

    Axial deformity secondary to degenerative joint disease of the knee can modify stress forces. Certain studies have reported an inversely proportional relationship between degenerative disease and osteoporosis. The aim of this prospective study was to quantify the horizontal linear distribution of bone density using dual x-ray absorptiometry (DXA) of the proximal tibia as a function of the femoral neck bone density in patients with knee osteoarthritis. Between September 1996 and March 1998, 90 cases of primary degenerative joint disease of the knee were programmed for total knee arthroplasty. Prior to the procedure, the patients were assessed clinically and radiologically according to the International Knee Society (IKS) criteria. The mechanical femorotibial angle was measured in all patients and the varus angles were recorded. Most of the patients were women (65 p. 100) with a mean age of 70 +/- 5 years. Valgus knees were excluded from this series. The mean mechanical femorotibial angle was 172 +/- 5 degrees. Fifteen patients had a normal axis (16 p. 100), 32 had a varus measuring 4 degrees to 10 degrees (35 p. 100) and 43 had a varus measuring 10 degrees or more (48 p. 100). The overall varus distance was 6.4 +/- 2 cm. All patients had two DXA explorations: femoral neck to determine the bone status according to the WHO criteria (normal, osteopenia, osteoporosis), knee to determine the linear distribution of bone density of the proximal tibia. A 7 mm high band including 7 regions of interest covering the width of the tibia were explored in the area where the tibial cut was to be made. These 7 regions of interest were: R1, R2 under the lateral compartment, R6, R7 under the medial compartment, and R3, R4, R5 on either side of the tibial spines. The level of significance was set at 5 p. 100. The mean Z score (0.54 +/- 1) in the 90 patients showed a symmetrical distribution. These patients were representative of their age range. Their T score was - 1.40 +/- 1 (m +/- SD) and most had osteopenia (54 p. 100) according to the WHO criteria, although 16 p. 100 had osteoporosis. Mean bone density of the knee was 0.898 +/- 0.163 g/cm(3) and was correlated with that of the femoral neck (r=0.61, p=0.001). There were significant correlations between the differences in the bone densities of the knee compartments (R6-R2, R7-R1) and the mechanical femorotibial angle [(r=0.39, p=0.0001); (r=0.52, p=0.001)]. Irrespective of the overall bone density, there was a strong medial compartment overloading, which correlated with the degree of varus deformation. DXA assessment of bone mineral density of the proximal tibia is a simple, reliable, precise and reproducible method. The distribution of bone density in the degenerative knee depends on the degree of deformation. The average level depends on the subject's general state of mineralization. Osteoporosis does not protect against degeneration of the knee joint since 16 p. 100 of our patients had osteoporosis according to the WHO criteria.

  20. Radiographic Evaluation of Ankle Joint Stability After Calcaneofibular Ligament Elevation During Open Reduction and Internal Fixation of Calcaneus Fracture.

    PubMed

    Wang, Chien-Shun; Tzeng, Yun-Hsuan; Lin, Chun-Cheng; Huang, Ching-Kuei; Chang, Ming-Chau; Chiang, Chao-Ching

    2016-09-01

    The aim of this prospective study was to evaluate the influence of sectioning the calcaneofibular ligament (CFL) during an extensile lateral approach during open reduction and internal fixation (ORIF) of calcaneal fractures on ankle joint stability. Forty-two patients with calcaneal fractures that received ORIF were included. Talar tilt stress and anterior drawer radiographs were performed on the operative and contralateral ankles 6 months postoperatively. The average degree of talar tilt on stress radiographs was 3.4 degrees (range, 0-12 degrees) on the operative side and 3.2 degrees (range, 0-14 degrees) on the contralateral side. The mean anterior drawer on stress radiographs of the CFL incised ankle was 6.1 mm (range, 2.4-11.8 mm) and on the contralateral ankle was 5.7 mm (range, 2.6-8.6 mm). There was no statistically significant difference of talar tilt and anterior drawer between the CFL incised side and the contralateral side (P = .658 and .302, respectively). The results suggest that sectioning of the CFL without any repair during ORIF of a calcaneal fracture does not have a negative effect on stability of the ankle. Repair of the CFL is, thus, probably not necessary following extended lateral approach for ORIF of calcaneal fractures. Level II, comparative study. © The Author(s) 2016.

Top