SU-F-T-469: A Clinically Observed Discrepancy Between Image-Based and Log- Based MLC Position
DOE Office of Scientific and Technical Information (OSTI.GOV)
Neal, B; Ahmed, M; Siebers, J
2016-06-15
Purpose: To present a clinical case which challenges the base assumption of log-file based QA, by showing that the actual position of a MLC leaf can suddenly deviate from its programmed and logged position by >1 mm as observed with real-time imaging. Methods: An EPID-based exit-fluence dosimetry system designed to prevent gross delivery errors was used in cine mode to capture portal images during treatment. Visual monitoring identified an anomalous MLC leaf pair gap not otherwise detected by the automatic position verification. The position of the erred leaf was measured on EPID images and log files were analyzed for themore » treatment in question, the prior day’s treatment, and for daily MLC test patterns acquired on those treatment days. Additional standard test patterns were used to quantify the leaf position. Results: Whereas the log file reported no difference between planned and recorded positions, image-based measurements showed the leaf to be 1.3±0.1 mm medial from the planned position. This offset was confirmed with the test pattern irradiations. Conclusion: It has been clinically observed that log-file derived leaf positions can differ from their actual positions by >1 mm, and therefore cannot be considered to be the actual leaf positions. This cautions the use of log-based methods for MLC or patient quality assurance without independent confirmation of log integrity. Frequent verification of MLC positions through independent means is a necessary precondition to trusting log file records. Intra-treatment EPID imaging provides a method to capture departures from MLC planned positions. Work was supported in part by Varian Medical Systems.« less
A clinically observed discrepancy between image-based and log-based MLC positions
DOE Office of Scientific and Technical Information (OSTI.GOV)
Neal, Brian, E-mail: bpn2p@virginia.edu; Ahmed, Mahmoud; Kathuria, Kunal
2016-06-15
Purpose: To present a clinical case in which real-time intratreatment imaging identified an multileaf collimator (MLC) leaf to be consistently deviating from its programmed and logged position by >1 mm. Methods: An EPID-based exit-fluence dosimetry system designed to prevent gross delivery errors was used to capture cine during treatment images. The author serendipitously visually identified a suspected MLC leaf displacement that was not otherwise detected. The leaf position as recorded on the EPID images was measured and log-files were analyzed for the treatment in question, the prior day’s treatment, and for daily MLC test patterns acquired on those treatment days.more » Additional standard test patterns were used to quantify the leaf position. Results: Whereas the log-file reported no difference between planned and recorded positions, image-based measurements showed the leaf to be 1.3 ± 0.1 mm medial from the planned position. This offset was confirmed with the test pattern irradiations. Conclusions: It has been clinically observed that log-file derived leaf positions can differ from their actual position by >1 mm, and therefore cannot be considered to be the actual leaf positions. This cautions the use of log-based methods for MLC or patient quality assurance without independent confirmation of log integrity. Frequent verification of MLC positions through independent means is a necessary precondition to trust log-file records. Intratreatment EPID imaging provides a method to capture departures from MLC planned positions.« less
SU-E-T-392: Evaluation of Ion Chamber/film and Log File Based QA to Detect Delivery Errors
DOE Office of Scientific and Technical Information (OSTI.GOV)
Nelson, C; Mason, B; Kirsner, S
2015-06-15
Purpose: Ion chamber and film (ICAF) is a method used to verify patient dose prior to treatment. More recently, log file based QA has been shown as an alternative for measurement based QA. In this study, we delivered VMAT plans with and without errors to determine if ICAF and/or log file based QA was able to detect the errors. Methods: Using two VMAT patients, the original treatment plan plus 7 additional plans with delivery errors introduced were generated and delivered. The erroneous plans had gantry, collimator, MLC, gantry and collimator, collimator and MLC, MLC and gantry, and gantry, collimator, andmore » MLC errors. The gantry and collimator errors were off by 4{sup 0} for one of the two arcs. The MLC error introduced was one in which the opening aperture didn’t move throughout the delivery of the field. For each delivery, an ICAF measurement was made as well as a dose comparison based upon log files. Passing criteria to evaluate the plans were ion chamber less and 5% and film 90% of pixels pass the 3mm/3% gamma analysis(GA). For log file analysis 90% of voxels pass the 3mm/3% 3D GA and beam parameters match what was in the plan. Results: Two original plans were delivered and passed both ICAF and log file base QA. Both ICAF and log file QA met the dosimetry criteria on 4 of the 12 erroneous cases analyzed (2 cases were not analyzed). For the log file analysis, all 12 erroneous plans alerted a mismatch in delivery versus what was planned. The 8 plans that didn’t meet criteria all had MLC errors. Conclusion: Our study demonstrates that log file based pre-treatment QA was able to detect small errors that may not be detected using an ICAF and both methods of were able to detect larger delivery errors.« less
SU-E-T-784: Using MLC Log Files for Daily IMRT Delivery Verification
DOE Office of Scientific and Technical Information (OSTI.GOV)
Stathakis, S; Defoor, D; Linden, P
2015-06-15
Purpose: To verify daily intensity modulated radiation therapy (IMRT) treatments using multi-leaf collimator (MLC) log files. Methods: The MLC log files from a NovalisTX Varian linear accelerator were used in this study. The MLC files were recorded daily for all patients undergoing IMRT or volumetric modulated arc therapy (VMAT). The first record of each patient was used as reference and all records for subsequent days were compared against the reference. An in house MATLAB software code was used for the comparisons. Each MLC log file was converted to a fluence map (FM) and a gamma index (γ) analysis was usedmore » for the evaluation of each daily delivery for every patient. The tolerance for the gamma index was set to 2% dose difference and 2mm distance to agreement while points with signal of 10% or lower of the maximum value were excluded from the comparisons. Results: The γ between each of the reference FMs and the consecutive daily fraction FMs had an average value of 99.1% (ranged from 98.2 to 100.0%). The FM images were reconstructed at various resolutions in order to study the effect of the resolution on the γ and at the same time reduce the time for processing the images. We found that the comparison of images with the highest resolution (768×1024) yielded on average a lower γ (99.1%) than the ones with low resolution (192×256) (γ 99.5%). Conclusion: We developed an in-house software that allows us to monitor the quality of daily IMRT and VMAT treatment deliveries using information from the MLC log files of the linear accelerator. The information can be analyzed and evaluated as early as after the completion of each daily treatment. Such tool can be valuable to assess the effect of MLC positioning on plan quality, especially in the context of adaptive radiotherapy.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Takahashi, R; Kamima, T; Tachibana, H
2016-06-15
Purpose: To investigate the effect of the trajectory files from linear accelerator for Clarkson-based independent dose verification in IMRT and VMAT plans. Methods: A CT-based independent dose verification software (Simple MU Analysis: SMU, Triangle Products, Japan) with a Clarksonbased algorithm was modified to calculate dose using the trajectory log files. Eclipse with the three techniques of step and shoot (SS), sliding window (SW) and Rapid Arc (RA) was used as treatment planning system (TPS). In this study, clinically approved IMRT and VMAT plans for prostate and head and neck (HN) at two institutions were retrospectively analyzed to assess the dosemore » deviation between DICOM-RT plan (PL) and trajectory log file (TJ). An additional analysis was performed to evaluate MLC error detection capability of SMU when the trajectory log files was modified by adding systematic errors (0.2, 0.5, 1.0 mm) and random errors (5, 10, 30 mm) to actual MLC position. Results: The dose deviations for prostate and HN in the two sites were 0.0% and 0.0% in SS, 0.1±0.0%, 0.1±0.1% in SW and 0.6±0.5%, 0.7±0.9% in RA, respectively. The MLC error detection capability shows the plans for HN IMRT were the most sensitive and 0.2 mm of systematic error affected 0.7% dose deviation on average. Effect of the MLC random error did not affect dose error. Conclusion: The use of trajectory log files including actual information of MLC location, gantry angle, etc should be more effective for an independent verification. The tolerance level for the secondary check using the trajectory file may be similar to that of the verification using DICOM-RT plan file. From the view of the resolution of MLC positional error detection, the secondary check could detect the MLC position error corresponding to the treatment sites and techniques. This research is partially supported by Japan Agency for Medical Research and Development (AMED)« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Stathakis, S; Defoor, D; Linden, P
Purpose: To study the frequency of Multi-Leaf Collimator (MLC) leaf failures, investigate methods to predict them and reduce linac downtime. Methods: A Varian HD120 MLC was used in our study. The hyperterminal MLC errors logged from 06/2012 to 12/2014 were collected. Along with the hyperterminal errors, the MLC motor changes and all other MLC interventions by the linear accelerator engineer were recorded. The MLC dynalog files were also recorded on a daily basis for each treatment and during linac QA. The dynalog files were analyzed to calculate root mean square errors (RMS) and cumulative MLC travel distance per motor. Anmore » in-house MatLab code was used to analyze all dynalog files, record RMS errors and calculate the distance each MLC traveled per day. Results: A total of 269 interventions were recorded over a period of 18 months. Of these, 146 included MLC motor leaf change, 39 T-nut replacements, and 84 MLC cleaning sessions. Leaves close to the middle of each side required the most maintenance. In the A bank, leaves A27 to A40 recorded 73% of all interventions, while the same leaves in the B bank counted for 52% of the interventions. On average, leaves in the middle of the bank had their motors changed approximately every 1500m of travel. Finally, it was found that the number of RMS errors increased prior to an MLC motor change. Conclusion: An MLC dynalog file analysis software was developed that can be used to log daily MLC usage. Our eighteen-month data analysis showed that there is a correlation between the distance an MLC travels, the RMS and the life of the MLC motor. We plan to use this tool to predict MLC motor failures and with proper and timely intervention, reduce the downtime of the linac during clinical hours.« less
SU-E-T-473: A Patient-Specific QC Paradigm Based On Trajectory Log Files and DICOM Plan Files
DOE Office of Scientific and Technical Information (OSTI.GOV)
DeMarco, J; McCloskey, S; Low, D
Purpose: To evaluate a remote QC tool for monitoring treatment machine parameters and treatment workflow. Methods: The Varian TrueBeamTM linear accelerator is a digital machine that records machine axis parameters and MLC leaf positions as a function of delivered monitor unit or control point. This information is saved to a binary trajectory log file for every treatment or imaging field in the patient treatment session. A MATLAB analysis routine was developed to parse the trajectory log files for a given patient, compare the expected versus actual machine and MLC positions as well as perform a cross-comparison with the DICOM-RT planmore » file exported from the treatment planning system. The parsing routine sorts the trajectory log files based on the time and date stamp and generates a sequential report file listing treatment parameters and provides a match relative to the DICOM-RT plan file. Results: The trajectory log parsing-routine was compared against a standard record and verify listing for patients undergoing initial IMRT dosimetry verification and weekly and final chart QC. The complete treatment course was independently verified for 10 patients of varying treatment site and a total of 1267 treatment fields were evaluated including pre-treatment imaging fields where applicable. In the context of IMRT plan verification, eight prostate SBRT plans with 4-arcs per plan were evaluated based on expected versus actual machine axis parameters. The average value for the maximum RMS MLC error was 0.067±0.001mm and 0.066±0.002mm for leaf bank A and B respectively. Conclusion: A real-time QC analysis program was tested using trajectory log files and DICOM-RT plan files. The parsing routine is efficient and able to evaluate all relevant machine axis parameters during a patient treatment course including MLC leaf positions and table positions at time of image acquisition and during treatment.« less
SU-F-T-233: Evaluation of Treatment Delivery Parameters Using High Resolution ELEKTA Log Files
DOE Office of Scientific and Technical Information (OSTI.GOV)
Kabat, C; Defoor, D; Alexandrian, A
2016-06-15
Purpose: As modern linacs have become more technologically advanced with the implementation of IGRT and IMRT with HDMLCs, a requirement for more elaborate tracking techniques to monitor components’ integrity is paramount. ElektaLog files are generated every 40 milliseconds, which can be analyzed to track subtle changes and provide another aspect of quality assurance. This allows for constant monitoring of fraction consistency in addition to machine reliability. With this in mind, it was the aim of the study to evaluate if ElektaLog files can be utilized for linac consistency QA. Methods: ElektaLogs were reviewed for 16 IMRT patient plans with >16more » fractions. Logs were analyzed by creating fluence maps from recorded values of MLC locations, jaw locations, and dose per unit time. Fluence maps were then utilized to calculate a 2D gamma index with a 2%–2mm criteria for each fraction. ElektaLogs were also used to analyze positional errors for MLC leaves and jaws, which were used to compute an overall error for the MLC banks, Y-jaws, and X-jaws by taking the root-meansquare value of the individual recorded errors during treatment. Additionally, beam on time was calculated using the number of ElektaLog file entries within the file. Results: The average 2D gamma for all 16 patient plans was found to be 98.0±2.0%. Recorded gamma index values showed an acceptable correlation between fractions. Average RMS values for MLC leaves and the jaws resulted in a leaf variation of roughly 0.3±0.08 mm and jaw variation of about 0.15±0.04 mm, both of which fall within clinical tolerances. Conclusion: The use of ElektaLog files for day-to-day evaluation of linac integrity and patient QA can be utilized to allow for reliable analysis of system accuracy and performance.« less
Automating linear accelerator quality assurance.
Eckhause, Tobias; Al-Hallaq, Hania; Ritter, Timothy; DeMarco, John; Farrey, Karl; Pawlicki, Todd; Kim, Gwe-Ya; Popple, Richard; Sharma, Vijeshwar; Perez, Mario; Park, SungYong; Booth, Jeremy T; Thorwarth, Ryan; Moran, Jean M
2015-10-01
The purpose of this study was 2-fold. One purpose was to develop an automated, streamlined quality assurance (QA) program for use by multiple centers. The second purpose was to evaluate machine performance over time for multiple centers using linear accelerator (Linac) log files and electronic portal images. The authors sought to evaluate variations in Linac performance to establish as a reference for other centers. The authors developed analytical software tools for a QA program using both log files and electronic portal imaging device (EPID) measurements. The first tool is a general analysis tool which can read and visually represent data in the log file. This tool, which can be used to automatically analyze patient treatment or QA log files, examines the files for Linac deviations which exceed thresholds. The second set of tools consists of a test suite of QA fields, a standard phantom, and software to collect information from the log files on deviations from the expected values. The test suite was designed to focus on the mechanical tests of the Linac to include jaw, MLC, and collimator positions during static, IMRT, and volumetric modulated arc therapy delivery. A consortium of eight institutions delivered the test suite at monthly or weekly intervals on each Linac using a standard phantom. The behavior of various components was analyzed for eight TrueBeam Linacs. For the EPID and trajectory log file analysis, all observed deviations which exceeded established thresholds for Linac behavior resulted in a beam hold off. In the absence of an interlock-triggering event, the maximum observed log file deviations between the expected and actual component positions (such as MLC leaves) varied from less than 1% to 26% of published tolerance thresholds. The maximum and standard deviations of the variations due to gantry sag, collimator angle, jaw position, and MLC positions are presented. Gantry sag among Linacs was 0.336 ± 0.072 mm. The standard deviation in MLC position, as determined by EPID measurements, across the consortium was 0.33 mm for IMRT fields. With respect to the log files, the deviations between expected and actual positions for parameters were small (<0.12 mm) for all Linacs. Considering both log files and EPID measurements, all parameters were well within published tolerance values. Variations in collimator angle, MLC position, and gantry sag were also evaluated for all Linacs. The performance of the TrueBeam Linac model was shown to be consistent based on automated analysis of trajectory log files and EPID images acquired during delivery of a standardized test suite. The results can be compared directly to tolerance thresholds. In addition, sharing of results from standard tests across institutions can facilitate the identification of QA process and Linac changes. These reference values are presented along with the standard deviation for common tests so that the test suite can be used by other centers to evaluate their Linac performance against those in this consortium.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Hirashima, H; Miyabe, Y; Yokota, K
2016-06-15
Purpose: The Dynamic Wave Arc (DWA) technique, where the multi-leaf collimator (MLC) and gantry/ring move simultaneously in a predefined non-coplanar trajectory, has been developed on the Vero4DRT. The aim of this study is to develop a simple method for quality assurance of DWA delivery using an electronic portal imaging device (EPID) measurements and log files analysis. Methods: The Vero4DRT has an EPID on the beam axis, the resolution of which is 0.18 mm/pixel at the isocenter plane. EPID images were acquired automatically. To verify the detection accuracy of the MLC position by EPID images, the MLC position with intentional errorsmore » was assessed. Tests were designed considering three factors: (1) accuracy of the MLC position (2) dose output consistency with variable dose rate (160–400 MU/min), gantry speed (2.4–6°/s), ring speed (0.5–2.5°/s), and (3) MLC speed (1.6–4.2 cm/s). All the patterns were delivered to the EPID and compared with those obtained with a stationary radiation beam with a 0° gantry angle. The irradiation log, including the MLC position and gantry/ring angle, were recorded simultaneously. To perform independent checks of the machine accuracy, the MLC position and gantry/ring angle position were assessed using log files. Results: 0.1 mm intentional error can be detected by the EPID, which is smaller than the EPID pixel size. The dose outputs with different conditions of the dose rate and gantry/ring speed and MLC speed showed good agreement, with a root mean square (RMS) error of 0.76%. The RMS error between the detected and recorded data were 0.1 mm for the MLC position, 0.12° for the gantry angle, and 0.07° for the ring angle. Conclusion: The MLC position and dose outputs in variable conditions during DWA irradiation can be easily detected using EPID measurements and log file analysis. The proposed method is useful for routine verification. This research is (partially) supported by the Practical Research for Innovative Cancer Control (15Ack0106151h0001) from Japan Agency for Medical Research and development, AMED. Authors Takashi Mizowaki and Masahiro Hiraoka have consultancy agreement with Mitsubishi Heavy Industries, Ltd., Japan.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Able, CM; Baydush, AH; Nguyen, C
Purpose: To determine the effectiveness of SPC analysis for a model predictive maintenance process that uses accelerator generated parameter and performance data contained in trajectory log files. Methods: Each trajectory file is decoded and a total of 131 axes positions are recorded (collimator jaw position, gantry angle, each MLC, etc.). This raw data is processed and either axis positions are extracted at critical points during the delivery or positional change over time is used to determine axis velocity. The focus of our analysis is the accuracy, reproducibility and fidelity of each axis. A reference positional trace of the gantry andmore » each MLC is used as a motion baseline for cross correlation (CC) analysis. A total of 494 parameters (482 MLC related) were analyzed using Individual and Moving Range (I/MR) charts. The chart limits were calculated using a hybrid technique that included the use of the standard 3σ limits and parameter/system specifications. Synthetic errors/changes were introduced to determine the initial effectiveness of I/MR charts in detecting relevant changes in operating parameters. The magnitude of the synthetic errors/changes was based on: TG-142 and published analysis of VMAT delivery accuracy. Results: All errors introduced were detected. Synthetic positional errors of 2mm for collimator jaw and MLC carriage exceeded the chart limits. Gantry speed and each MLC speed are analyzed at two different points in the delivery. Simulated Gantry speed error (0.2 deg/sec) and MLC speed error (0.1 cm/sec) exceeded the speed chart limits. Gantry position error of 0.2 deg was detected by the CC maximum value charts. The MLC position error of 0.1 cm was detected by the CC maximum value location charts for every MLC. Conclusion: SPC I/MR evaluation of trajectory log file parameters may be effective in providing an early warning of performance degradation or component failure for medical accelerator systems.« less
SU-E-T-261: Plan Quality Assurance of VMAT Using Fluence Images Reconstituted From Log-Files
DOE Office of Scientific and Technical Information (OSTI.GOV)
Katsuta, Y; Shimizu, E; Matsunaga, K
2014-06-01
Purpose: A successful VMAT plan delivery includes precise modulations of dose rate, gantry rotational and multi-leaf collimator (MLC) shapes. One of the main problem in the plan quality assurance is dosimetric errors associated with leaf-positional errors are difficult to analyze because they vary with MU delivered and leaf number. In this study, we calculated integrated fluence error image (IFEI) from log-files and evaluated plan quality in the area of all and individual MLC leaves scanned. Methods: The log-file reported the expected and actual position for inner 20 MLC leaves and the dose fraction every 0.25 seconds during prostate VMAT onmore » Elekta Synergy. These data were imported to in-house software that developed to calculate expected and actual fluence images from the difference of opposing leaf trajectories and dose fraction at each time. The IFEI was obtained by adding all of the absolute value of the difference between expected and actual fluence images corresponding. Results: In the area all MLC leaves scanned in the IFEI, the average and root mean square (rms) were 2.5 and 3.6 MU, the area of errors below 10, 5 and 3 MU were 98.5, 86.7 and 68.1 %, the 95 % of area was covered with less than error of 7.1 MU. In the area individual MLC leaves scanned in the IFEI, the average and rms value were 2.1 – 3.0 and 3.1 – 4.0 MU, the area of errors below 10, 5 and 3 MU were 97.6 – 99.5, 81.7 – 89.5 and 51.2 – 72.8 %, the 95 % of area was covered with less than error of 6.6 – 8.2 MU. Conclusion: The analysis of the IFEI reconstituted from log-file was provided detailed information about the delivery in the area of all and individual MLC leaves scanned.« less
Ranade, Manisha K; Lynch, Bart D; Li, Jonathan G; Dempsey, James F
2006-01-01
We have developed an electronic portal imaging device (EPID) employing a fast scintillator and a high-speed camera. The device is designed to accurately and independently characterize the fluence delivered by a linear accelerator during intensity modulated radiation therapy (IMRT) with either step-and-shoot or dynamic multileaf collimator (MLC) delivery. Our aim is to accurately obtain the beam shape and fluence of all segments delivered during IMRT, in order to study the nature of discrepancies between the plan and the delivered doses. A commercial high-speed camera was combined with a terbium-doped gadolinium-oxy-sulfide (Gd2O2S:Tb) scintillator to form an EPID for the unaliased capture of two-dimensional fluence distributions of each beam in an IMRT delivery. The high speed EPID was synchronized to the accelerator pulse-forming network and gated to capture every possible pulse emitted from the accelerator, with an approximate frame rate of 360 frames-per-second (fps). A 62-segment beam from a head-and-neck IMRT treatment plan requiring 68 s to deliver was recorded with our high speed EPID producing approximately 6 Gbytes of imaging data. The EPID data were compared with the MLC instruction files and the MLC controller log files. The frames were binned to provide a frame rate of 72 fps with a signal-to-noise ratio that was sufficient to resolve leaf positions and segment fluence. The fractional fluence from the log files and EPID data agreed well. An ambiguity in the motion of the MLC during beam on was resolved. The log files reported leaf motions at the end of 33 of the 42 segments, while the EPID observed leaf motions in only 7 of the 42 segments. The static IMRT segment shapes observed by the high speed EPID were in good agreement with the shapes reported in the log files. The leaf motions observed during beam-on for step-and-shoot delivery were not temporally resolved by the log files.
[Investigation of Elekta linac characteristics for VMAT].
Luo, Guangwen; Zhang, Kunyi
2012-01-01
The aim of this study is to investigate the characteristics of Elekta delivery system for volumetric modulated arc therapy (VMAT). Five VMAT plans were delivered in service mode and dose rates, and speed of gantry and MLC leaves were analyzed by log files. Results showed that dose rates varied between 6 dose rates. Gantry and MLC leaf speed dynamically varied during delivery. The technique of VMAT requires linac to dynamically control more parameters, and these key dynamic variables during VMAT delivery can be checked by log files. Quality assurance procedure should be carried out for VMAT related parameter.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Tateoka, K; Graduate School of Medicine, Sapporo Medical University, Sapporo, JP; Fujimomo, K
2014-06-01
Purpose: The aim of the study is to evaluate the use of Varian DynaLog files to verify VMAT plans delivery and modulation complexity score (MCS) of VMAT. Methods: Delivery accuracy of machine performance was quantified by multileaf collimator (MLC) position errors, gantry angle errors and fluence delivery accuracy for volumetric modulated arc therapy (VMAT). The relationship between machine performance and plan complexity were also investigated using the modulation complexity score (MCS). Plan and Actual MLC positions, gantry angles and delivered fraction of monitor units were extracted from Varian DynaLog files. These factors were taken from the record and verify systemmore » of MLC control file. Planned and delivered beam data were compared to determine leaf position errors and gantry angle errors. Analysis was also performed on planned and actual fluence maps reconstructed from those of the DynaLog files. This analysis was performed for all treatment fractions of 5 prostate VMAT plans. The analysis of DynaLog files have been carried out by in-house programming in Visual C++. Results: The root mean square of leaf position and gantry angle errors were about 0.12 and 0.15, respectively. The Gamma of planned and actual fluence maps at 3%/3 mm criterion was about 99.21. The gamma of the leaf position errors were not directly related to plan complexity as determined by the MCS. Therefore, the gamma of the gantry angle errors were directly related to plan complexity as determined by the MCS. Conclusion: This study shows Varian dynalog files for VMAT plan can be diagnosed delivery errors not possible with phantom based quality assurance. Furthermore, the MCS of VMAT plan can evaluate delivery accuracy for patients receiving of VMAT. Machine performance was found to be directly related to plan complexity but this is not the dominant determinant of delivery accuracy.« less
Pasler, Marlies; Kaas, Jochem; Perik, Thijs; Geuze, Job; Dreindl, Ralf; Künzler, Thomas; Wittkamper, Frits; Georg, Dietmar
2015-12-01
To systematically evaluate machine specific quality assurance (QA) for volumetric modulated arc therapy (VMAT) based on log files by applying a dynamic benchmark plan. A VMAT benchmark plan was created and tested on 18 Elekta linacs (13 MLCi or MLCi2, 5 Agility) at 4 different institutions. Linac log files were analyzed and a delivery robustness index was introduced. For dosimetric measurements an ionization chamber array was used. Relative dose deviations were assessed by mean gamma for each control point and compared to the log file evaluation. Fourteen linacs delivered the VMAT benchmark plan, while 4 linacs failed by consistently terminating the delivery. The mean leaf error (±1SD) was 0.3±0.2 mm for all linacs. Large MLC maximum errors up to 6.5 mm were observed at reversal positions. Delivery robustness index accounting for MLC position correction (0.8-1.0) correlated with delivery time (80-128 s) and depended on dose rate performance. Dosimetric evaluation indicated in general accurate plan reproducibility with γ(mean)(±1 SD)=0.4±0.2 for 1 mm/1%. However single control point analysis revealed larger deviations and attributed well to log file analysis. The designed benchmark plan helped identify linac related malfunctions in dynamic mode for VMAT. Log files serve as an important additional QA measure to understand and visualize dynamic linac parameters. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.
Quantification of residual dose estimation error on log file-based patient dose calculation.
Katsuta, Yoshiyuki; Kadoya, Noriyuki; Fujita, Yukio; Shimizu, Eiji; Matsunaga, Kenichi; Matsushita, Haruo; Majima, Kazuhiro; Jingu, Keiichi
2016-05-01
The log file-based patient dose estimation includes a residual dose estimation error caused by leaf miscalibration, which cannot be reflected on the estimated dose. The purpose of this study is to determine this residual dose estimation error. Modified log files for seven head-and-neck and prostate volumetric modulated arc therapy (VMAT) plans simulating leaf miscalibration were generated by shifting both leaf banks (systematic leaf gap errors: ±2.0, ±1.0, and ±0.5mm in opposite directions and systematic leaf shifts: ±1.0mm in the same direction) using MATLAB-based (MathWorks, Natick, MA) in-house software. The generated modified and non-modified log files were imported back into the treatment planning system and recalculated. Subsequently, the generalized equivalent uniform dose (gEUD) was quantified for the definition of the planning target volume (PTV) and organs at risks. For MLC leaves calibrated within ±0.5mm, the quantified residual dose estimation errors that obtained from the slope of the linear regression of gEUD changes between non- and modified log file doses per leaf gap are in head-and-neck plans 1.32±0.27% and 0.82±0.17Gy for PTV and spinal cord, respectively, and in prostate plans 1.22±0.36%, 0.95±0.14Gy, and 0.45±0.08Gy for PTV, rectum, and bladder, respectively. In this work, we determine the residual dose estimation errors for VMAT delivery using the log file-based patient dose calculation according to the MLC calibration accuracy. Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
SU-F-T-295: MLCs Performance and Patient-Specific IMRT QA Using Log File Analysis
DOE Office of Scientific and Technical Information (OSTI.GOV)
Osman, A; American University of Biuret Medical Center, Biuret; Maalej, N
2016-06-15
Purpose: To analyze the performance of the multi-leaf collimators (MLCs) from the log files recorded during the intensity modulated radiotherapy (IMRT) treatment and to construct the relative fluence maps and do the gamma analysis to compare the planned and executed MLCs movement. Methods: We developed a program to extract and analyze the data from dynamic log files (dynalog files) generated from sliding window IMRT delivery treatments. The program extracts the planned and executed (actual or delivered) MLCs movement, calculates and compares the relative planned and executed fluences. The fluence maps were used to perform the gamma analysis (with 3% dosemore » difference and 3 mm distance to agreement) for 3 IMR patients. We compared our gamma analysis results with those obtained from portal dose image prediction (PDIP) algorithm performed using the EPID. Results: For 3 different IMRT patient treatments, the maximum difference between the planned and the executed MCLs positions was 1.2 mm. The gamma analysis results of the planned and delivered fluences were in good agreement with the gamma analysis from portal dosimetry. The maximum difference for number of pixels passing the gamma criteria (3%/3mm) was 0.19% with respect to portal dosimetry results. Conclusion: MLC log files can be used to verify the performance of the MLCs. Patientspecific IMRT QA based on MLC movement log files gives similar results to EPID dosimetry results. This promising method for patient-specific IMRT QA is fast, does not require dose measurements in a phantom, can be done before the treatment and for every fraction, and significantly reduces the IMRT workload. The author would like to thank King Fahd University of petroleum and Minerals for the support.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Morcos, M; Mitrou, E
2015-06-15
Purpose: Using Linac dynamic logs (Dynalogs) we evaluate the impact of a single failing MLC motor on the deliverability of an IMRT plan by assessing the recalculated dose volume histograms (DVHs) taking the delivered MLC positions and beam hold-offs into consideration. Methods: This is a retrospective study based on a deteriorating MLC motor (leaf 36B) which was observed to be failing via Dynalog analysis. To investigate further, Eclipse-importable MLC files were generated from Dynalogs to recalculate the actual delivered dose and to assess the clinical impact through DVHs. All deliveries were performed on a Varian 21EX linear accelerator equipped withmore » Millennium-120 MLC. The analysis of Dynalog files and subsequent conversion to Eclipse-importable MLC files were all performed by in-house programming in Python. Effects on plan DVH are presented in the following section on a particular brain-IMRT plan which was delivered with a failing MLC motor which was then replaced. Results: Global max dose increased by 13.5%, max dose to the brainstem PRV increased by 8.2%, max dose to the optic chiasm increased by 7.6%, max dose to optic nerve increased by 8.8% and the mean dose to the PTV increased by 7.9% when comparing the original plan to the fraction with the failing MLC motor. The reason the dose increased was due to the failure being on the B-bank which is the lagging side on a sliding window delivery, therefore any failures on this side will cause an over-irradiation as the B-bank leaves struggles to keep the window from growing. Conclusion: Our findings suggest that a single failing MLC motor may jeopardize the entire delivery. This may be due to the bad MLC motor drawing too much current causing all MLCs on the same bank to underperform. This hypothesis will be investigated in a future study.« less
Monte Carlo based, patient-specific RapidArc QA using Linac log files.
Teke, Tony; Bergman, Alanah M; Kwa, William; Gill, Bradford; Duzenli, Cheryl; Popescu, I Antoniu
2010-01-01
A Monte Carlo (MC) based QA process to validate the dynamic beam delivery accuracy for Varian RapidArc (Varian Medical Systems, Palo Alto, CA) using Linac delivery log files (DynaLog) is presented. Using DynaLog file analysis and MC simulations, the goal of this article is to (a) confirm that adequate sampling is used in the RapidArc optimization algorithm (177 static gantry angles) and (b) to assess the physical machine performance [gantry angle and monitor unit (MU) delivery accuracy]. Ten clinically acceptable RapidArc treatment plans were generated for various tumor sites and delivered to a water-equivalent cylindrical phantom on the treatment unit. Three Monte Carlo simulations were performed to calculate dose to the CT phantom image set: (a) One using a series of static gantry angles defined by 177 control points with treatment planning system (TPS) MLC control files (planning files), (b) one using continuous gantry rotation with TPS generated MLC control files, and (c) one using continuous gantry rotation with actual Linac delivery log files. Monte Carlo simulated dose distributions are compared to both ionization chamber point measurements and with RapidArc TPS calculated doses. The 3D dose distributions were compared using a 3D gamma-factor analysis, employing a 3%/3 mm distance-to-agreement criterion. The dose difference between MC simulations, TPS, and ionization chamber point measurements was less than 2.1%. For all plans, the MC calculated 3D dose distributions agreed well with the TPS calculated doses (gamma-factor values were less than 1 for more than 95% of the points considered). Machine performance QA was supplemented with an extensive DynaLog file analysis. A DynaLog file analysis showed that leaf position errors were less than 1 mm for 94% of the time and there were no leaf errors greater than 2.5 mm. The mean standard deviation in MU and gantry angle were 0.052 MU and 0.355 degrees, respectively, for the ten cases analyzed. The accuracy and flexibility of the Monte Carlo based RapidArc QA system were demonstrated. Good machine performance and accurate dose distribution delivery of RapidArc plans were observed. The sampling used in the TPS optimization algorithm was found to be adequate.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Li, X; Yang, F
Purpose: Knowing MLC leaf positioning error over the course of treatment would be valuable for treatment planning, QA design, and patient safety. The objective of the current study was to quantify the MLC positioning accuracy for VMAT delivery of head and neck treatment plans. Methods: A total of 837 MLC log files were collected from 14 head and neck cancer patients undergoing full arc VMAT treatment on one Varian Trilogy machine. The actual and planned leaf gaps were extracted from the retrieved MLC log files. For a given patient, the leaf gap error percentage (LGEP), defined as the ratio ofmore » the actual leaf gap over the planned, was evaluated for each leaf pair at all the gantry angles recorded over the course of the treatment. Statistics describing the distribution of the largest LGEP (LLGEP) of the 60 leaf pairs including the maximum, minimum, mean, Kurtosis, and skewness were evaluated. Results: For the 14 studied patients, their PTV located at tonsil, base of tongue, larynx, supraglottis, nasal cavity, and thyroid gland with volume ranging from 72.0 cm{sup 3} to 602.0 cm{sup 3}. The identified LLGEP differed between patients. It ranged from 183.9% to 457.7% with a mean of 368.6%. For the majority of the patients, the LLGEP distributions peaked at non-zero positions and showed no obvious dependence on gantry rotations. Kurtosis and skewness, with minimum/maximum of 66.6/217.9 and 6.5/12.6, respectively, suggested relatively more peaked while right-skewed leaf error distribution pattern. Conclusion: The results indicate pattern of MLC leaf gap error differs between patients of lesion located at similar anatomic site. Understanding the systemic mechanisms underlying these observed error patterns necessitates examining more patient-specific plan parameters in a large patient cohort setting.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Calvo Ortega, Juan Francisco, E-mail: jfcdrr@yahoo.es; Moragues, Sandra; Pozo, Miquel
2014-07-01
The aim of this study is to assess the accuracy of a convolution-based algorithm (anisotropic analytical algorithm [AAA]) implemented in the Eclipse planning system for intensity-modulated radiosurgery (IMRS) planning of small cranial targets by using a 5-mm leaf-width multileaf collimator (MLC). Overall, 24 patient-based IMRS plans for cranial lesions of variable size (0.3 to 15.1 cc) were planned (Eclipse, AAA, version 10.0.28) using fixed field-based IMRS produced by a Varian linear accelerator equipped with a 120 MLC (5-mm width on central leaves). Plan accuracy was evaluated according to phantom-based measurements performed with radiochromic film (EBT2, ISP, Wayne, NJ). Film 2Dmore » dose distributions were performed with the FilmQA Pro software (version 2011, Ashland, OH) by using the triple-channel dosimetry method. Comparison between computed and measured 2D dose distributions was performed using the gamma method (3%/1 mm). Performance of the MLC was checked by inspection of the DynaLog files created by the linear accelerator during the delivery of each dynamic field. The absolute difference between the calculated and measured isocenter doses for all the IMRS plans was 2.5% ± 2.1%. The gamma evaluation method resulted in high average passing rates of 98.9% ± 1.4% (red channel) and 98.9% ± 1.5% (blue and green channels). DynaLog file analysis revealed a maximum root mean square error of 0.46 mm. According to our results, we conclude that the Eclipse/AAA algorithm provides accurate cranial IMRS dose distributions that may be accurately delivered by a Varian linac equipped with a Millennium 120 MLC.« less
The effect of MLC speed and acceleration on the plan delivery accuracy of VMAT
Park, J M; Wu, H-G; Kim, J H; Carlson, J N K
2015-01-01
Objective: To determine a new metric utilizing multileaf collimator (MLC) speeds and accelerations to predict plan delivery accuracy of volumetric modulated arc therapy (VMAT). Methods: To verify VMAT delivery accuracy, gamma evaluations, analysis of mechanical parameter difference between plans and log files, and analysis of changes in dose–volumetric parameters between plans and plans reconstructed with log files were performed with 40 VMAT plans. The average proportion of leaf speeds ranging from l to h cm s−1 (Sl–h and l–h = 0–0.4, 0.4–0.8, 0.8–1.2, 1.2–1.6 and 1.6–2.0), mean and standard deviation of MLC speeds were calculated for each VMAT plan. The same was carried out for accelerations in centimetre per second squared (Al–h and l–h = 0–4, 4–8, 8–12, 12–16 and 16–20). The correlations of those indicators to plan delivery accuracy were analysed with Spearman's correlation coefficient (rs). Results: The S1.2–1.6 and mean acceleration of MLCs showed generally higher correlations to plan delivery accuracy than did others. The highest rs values were observed between S1.2–1.6 and global 1%/2 mm (rs = −0.698 with p < 0.001) as well as mean acceleration and global 1%/2 mm (rs = −0.650 with p < 0.001). As the proportion of MLC speeds and accelerations >0.4 and 4 cm s−2 increased, the plan delivery accuracy of VMAT decreased. Conclusion: The variations in MLC speeds and accelerations showed considerable correlations to VMAT delivery accuracy. Advances in knowledge: As the MLC speeds and accelerations increased, VMAT delivery accuracy reduced. PMID:25734490
SU-F-T-465: Two Years of Radiotherapy Treatments Analyzed Through MLC Log Files
DOE Office of Scientific and Technical Information (OSTI.GOV)
Defoor, D; Kabat, C; Papanikolaou, N
Purpose: To present treatment statistics of a Varian Novalis Tx using more than 90,000 Varian Dynalog files collected over the past 2 years. Methods: Varian Dynalog files are recorded for every patient treated on our Varian Novalis Tx. The files are collected and analyzed daily to check interfraction agreement of treatment deliveries. This is accomplished by creating fluence maps from the data contained in the Dynalog files. From the Dynalog files we have also compiled statistics for treatment delivery times, MLC errors, gantry errors and collimator errors. Results: The mean treatment time for VMAT patients was 153 ± 86 secondsmore » while the mean treatment time for step & shoot was 256 ± 149 seconds. Patient’s treatment times showed a variation of 0.4% over there treatment course for VMAT and 0.5% for step & shoot. The average field sizes were 40 cm2 and 26 cm2 for VMAT and step & shoot respectively. VMAT beams contained and average overall leaf travel of 34.17 meters and step & shoot beams averaged less than half of that at 15.93 meters. When comparing planned and delivered fluence maps generated using the Dynalog files VMAT plans showed an average gamma passing percentage of 99.85 ± 0.47. Step & shoot plans showed an average gamma passing percentage of 97.04 ± 0.04. 5.3% of beams contained an MLC error greater than 1 mm and 2.4% had an error greater than 2mm. The mean gantry speed for VMAT plans was 1.01 degrees/s with a maximum of 6.5 degrees/s. Conclusion: Varian Dynalog files are useful for monitoring machine performance treatment parameters. The Dynalog files have shown that the performance of the Novalis Tx is consistent over the course of a patients treatment with only slight variations in patient treatment times and a low rate of MLC errors.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Wijesooriya, K; Seitter, K; Desai, V
Purpose: To present our single institution experience on catching errors with trajectory log file analysis. The reported causes of failures, probability of occurrences (O), severity of effects (S), and the probability of the failures to be undetected (D) could be added to guidelines of FMEA analysis. Methods: From March 2013 to March 2014, 19569 patient treatment fields/arcs were analyzed. This work includes checking all 131 treatment delivery parameters for all patients, all treatment sites and all treatment delivery fractions. TrueBeam trajectory log files for all treatment field types as well as all imaging types were accessed, read in every 20ms,more » and every control point (total of 37 million parameters) compared to the physician approved plan in the planning system. Results: Couch angle outlier occurrence: N= 327, range = −1.7 −1.2 deg; gantry angle outlier occurrence: N =59, range = 0.09 – 5.61 deg, collimator angle outlier occurrence: N = 13, range = −0.2 – 0.2 deg. VMAT cases have slightly larger variations in mechanical parameters. MLC: 3D single control point fields have a maximum deviation of 0.04 mm, 39 step and shoot IMRT cases have MLC −0.3 – 0.5 mm deviations, all (1286) VMAT cases have −0.9 – 0.7 mm deviations. Two possible serious errors were found: 1) A 4 cm isocenter shift for the PA beam of an AP-PA pair, under-dosing a portion of PTV by 25%. 2) Delivery with MLC leaves abutted behind the jaws as opposed to the midline as planned, leading to a under-dosing of a small volume of the PTV by 25%, by just the boost plan. Due to their error origin, neither of these errors could have been detected by pre-treatment verification. Conclusion: Performing Trajectory Log file analysis could catch typically undetected errors to avoid potentially adverse incidents.« less
A Varian DynaLog file-based procedure for patient dose-volume histogram-based IMRT QA.
Calvo-Ortega, Juan F; Teke, Tony; Moragues, Sandra; Pozo, Miquel; Casals-Farran, Joan
2014-03-06
In the present study, we describe a method based on the analysis of the dynamic MLC log files (DynaLog) generated by the controller of a Varian linear accelerator in order to perform patient-specific IMRT QA. The DynaLog files of a Varian Millennium MLC, recorded during an IMRT treatment, can be processed using a MATLAB-based code in order to generate the actual fluence for each beam and so recalculate the actual patient dose distribution using the Eclipse treatment planning system. The accuracy of the DynaLog-based dose reconstruction procedure was assessed by introducing ten intended errors to perturb the fluence of the beams of a reference plan such that ten subsequent erroneous plans were generated. In-phantom measurements with an ionization chamber (ion chamber) and planar dose measurements using an EPID system were performed to investigate the correlation between the measured dose changes and the expected ones detected by the reconstructed plans for the ten intended erroneous cases. Moreover, the method was applied to 20 cases of clinical plans for different locations (prostate, lung, breast, and head and neck). A dose-volume histogram (DVH) metric was used to evaluate the impact of the delivery errors in terms of dose to the patient. The ionometric measurements revealed a significant positive correlation (R² = 0.9993) between the variations of the dose induced in the erroneous plans with respect to the reference plan and the corresponding changes indicated by the DynaLog-based reconstructed plans. The EPID measurements showed that the accuracy of the DynaLog-based method to reconstruct the beam fluence was comparable with the dosimetric resolution of the portal dosimetry used in this work (3%/3 mm). The DynaLog-based reconstruction method described in this study is a suitable tool to perform a patient-specific IMRT QA. This method allows us to perform patient-specific IMRT QA by evaluating the result based on the DVH metric of the planning CT image (patient DVH-based IMRT QA).
Catching errors with patient-specific pretreatment machine log file analysis.
Rangaraj, Dharanipathy; Zhu, Mingyao; Yang, Deshan; Palaniswaamy, Geethpriya; Yaddanapudi, Sridhar; Wooten, Omar H; Brame, Scott; Mutic, Sasa
2013-01-01
A robust, efficient, and reliable quality assurance (QA) process is highly desired for modern external beam radiation therapy treatments. Here, we report the results of a semiautomatic, pretreatment, patient-specific QA process based on dynamic machine log file analysis clinically implemented for intensity modulated radiation therapy (IMRT) treatments delivered by high energy linear accelerators (Varian 2100/2300 EX, Trilogy, iX-D, Varian Medical Systems Inc, Palo Alto, CA). The multileaf collimator machine (MLC) log files are called Dynalog by Varian. Using an in-house developed computer program called "Dynalog QA," we automatically compare the beam delivery parameters in the log files that are generated during pretreatment point dose verification measurements, with the treatment plan to determine any discrepancies in IMRT deliveries. Fluence maps are constructed and compared between the delivered and planned beams. Since clinical introduction in June 2009, 912 machine log file analyses QA were performed by the end of 2010. Among these, 14 errors causing dosimetric deviation were detected and required further investigation and intervention. These errors were the result of human operating mistakes, flawed treatment planning, and data modification during plan file transfer. Minor errors were also reported in 174 other log file analyses, some of which stemmed from false positives and unreliable results; the origins of these are discussed herein. It has been demonstrated that the machine log file analysis is a robust, efficient, and reliable QA process capable of detecting errors originating from human mistakes, flawed planning, and data transfer problems. The possibility of detecting these errors is low using point and planar dosimetric measurements. Copyright © 2013 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Yang, D; Li, X; Li, H
2014-06-15
Purpose: Two aims of this work were to develop a method to automatically verify treatment delivery accuracy immediately after patient treatment and to develop a comprehensive daily treatment report to provide all required information for daily MR-IGRT review. Methods: After systematically analyzing the requirements for treatment delivery verification and understanding the available information from a novel MR-IGRT treatment machine, we designed a method to use 1) treatment plan files, 2) delivery log files, and 3) dosimetric calibration information to verify the accuracy and completeness of daily treatment deliveries. The method verifies the correctness of delivered treatment plans and beams, beammore » segments, and for each segment, the beam-on time and MLC leaf positions. Composite primary fluence maps are calculated from the MLC leaf positions and the beam-on time. Error statistics are calculated on the fluence difference maps between the plan and the delivery. We also designed the daily treatment delivery report by including all required information for MR-IGRT and physics weekly review - the plan and treatment fraction information, dose verification information, daily patient setup screen captures, and the treatment delivery verification results. Results: The parameters in the log files (e.g. MLC positions) were independently verified and deemed accurate and trustable. A computer program was developed to implement the automatic delivery verification and daily report generation. The program was tested and clinically commissioned with sufficient IMRT and 3D treatment delivery data. The final version has been integrated into a commercial MR-IGRT treatment delivery system. Conclusion: A method was developed to automatically verify MR-IGRT treatment deliveries and generate daily treatment reports. Already in clinical use since December 2013, the system is able to facilitate delivery error detection, and expedite physician daily IGRT review and physicist weekly chart review.« less
Zwan, Benjamin J; Barnes, Michael P; Hindmarsh, Jonathan; Lim, Seng B; Lovelock, Dale M; Fuangrod, Todsaporn; O'Connor, Daryl J; Keall, Paul J; Greer, Peter B
2017-08-01
An ideal commissioning and quality assurance (QA) program for Volumetric Modulated Arc Therapy (VMAT) delivery systems should assess the performance of each individual dynamic component as a function of gantry angle. Procedures within such a program should also be time-efficient, independent of the delivery system and be sensitive to all types of errors. The purpose of this work is to develop a system for automated time-resolved commissioning and QA of VMAT control systems which meets these criteria. The procedures developed within this work rely solely on images obtained, using an electronic portal imaging device (EPID) without the presence of a phantom. During the delivery of specially designed VMAT test plans, EPID frames were acquired at 9.5 Hz, using a frame grabber. The set of test plans was developed to individually assess the performance of the dose delivery and multileaf collimator (MLC) control systems under varying levels of delivery complexities. An in-house software tool was developed to automatically extract features from the EPID images and evaluate the following characteristics as a function of gantry angle: dose delivery accuracy, dose rate constancy, beam profile constancy, gantry speed constancy, dynamic MLC positioning accuracy, MLC speed and acceleration constancy, and synchronization between gantry angle, MLC positioning and dose rate. Machine log files were also acquired during each delivery and subsequently compared to information extracted from EPID image frames. The largest difference between measured and planned dose at any gantry angle was 0.8% which correlated with rapid changes in dose rate and gantry speed. For all other test plans, the dose delivered was within 0.25% of the planned dose for all gantry angles. Profile constancy was not found to vary with gantry angle for tests where gantry speed and dose rate were constant, however, for tests with varying dose rate and gantry speed, segments with lower dose rate and higher gantry speed exhibited less profile stability. MLC positional accuracy was not observed to be dependent on the degree of interdigitation. MLC speed was measured for each individual leaf and slower leaf speeds were shown to be compensated for by lower dose rates. The test procedures were found to be sensitive to 1 mm systematic MLC errors, 1 mm random MLC errors, 0.4 mm MLC gap errors and synchronization errors between the MLC, dose rate and gantry angle controls systems of 1°. In general, parameters measured by both EPID and log files agreed with the plan, however, a greater average departure from the plan was evidenced by the EPID measurements. QA test plans and analysis methods have been developed to assess the performance of each dynamic component of VMAT deliveries individually and as a function of gantry angle. This methodology relies solely on time-resolved EPID imaging without the presence of a phantom and has been shown to be sensitive to a range of delivery errors. The procedures developed in this work are both comprehensive and time-efficient and can be used for streamlined commissioning and QA of VMAT delivery systems. © 2017 American Association of Physicists in Medicine.
Modulation indices for volumetric modulated arc therapy.
Park, Jong Min; Park, So-Yeon; Kim, Hyoungnyoun; Kim, Jin Ho; Carlson, Joel; Ye, Sung-Joon
2014-12-07
The aim of this study is to present a modulation index (MI) for volumetric modulated arc therapy (VMAT) based on the speed and acceleration analysis of modulating-parameters such as multi-leaf collimator (MLC) movements, gantry rotation and dose-rate, comprehensively. The performance of the presented MI (MIt) was evaluated with correlation analyses to the pre-treatment quality assurance (QA) results, differences in modulating-parameters between VMAT plans versus dynamic log files, and differences in dose-volumetric parameters between VMAT plans versus reconstructed plans using dynamic log files. For comparison, the same correlation analyses were performed for the previously suggested modulation complexity score (MCS(v)), leaf travel modulation complexity score (LTMCS) and MI by Li and Xing (MI Li&Xing). In the two-tailed unpaired parameter condition, p values were acquired. The Spearman's rho (r(s)) values of MIt, MCSv, LTMCS and MI Li&Xing to the local gamma passing rate with 2%/2 mm criterion were -0.658 (p < 0.001), 0.186 (p = 0.251), 0.312 (p = 0.05) and -0.455 (p = 0.003), respectively. The values of rs to the modulating-parameter (MLC positions) differences were 0.917, -0.635, -0.857 and 0.795, respectively (p < 0.001). For dose-volumetric parameters, MIt showed higher statistically significant correlations than the conventional MIs. The MIt showed good performance for the evaluation of the modulation-degree of VMAT plans.
SU-E-T-550: Modulation Index for VMAT
DOE Office of Scientific and Technical Information (OSTI.GOV)
Park, J; Park, S; Kim, J
2015-06-15
Purpose: To present modulation indices (MIs) for volumetric modulated arc therapy (VMAT). Methods: A total of 40 VMAT plans were retrospectively selected. To investigate the delivery accuracy of each VMAT plan, gamma passing rates, differences in modulating parameters between plans and log files, and differences between the original plans and the plans reconstructed with the log files were acquired. A modulation index (MIt) was designed by multiplications of the weighted quantifications of MLC speeds, MLC accelerations, gantry accelerations and dose-rate variations. Textural features including angular second moment, inverse difference moment, contrast, variance, correlation and entropy were calculated from the fluencesmore » of each VMAT plan. To test the performance of suggested MIs, Spearman’s rank correlation coefficients (r) with the plan delivery accuracy were calculated. Conventional modulation indices for VMAT including the modulation complexity score for VMAT (MCSv), leaf travel modulation complexity score (LTMCS) and MI by Li & Xing were calculated, and their correlations were also analyzed in the same way. Results: The r values of contrast (particular displacement distance, d = 1), variance (d = 1), MIt, MCSv, LTMCS and MI by Li&Xing to the local gamma passing rates with 2%/2 mm were 0.547 (p < 0.001), 0.519 (p < 0.001), −0.658 (p < 0.001), 0.186 (p = 0.251), 0.312 (p = 0.05) and −0.455 (p = 0.003), respectively. The r values of those to the MLC errors were −0.863, −0.828, 0.917, −0.635, − 0.857 and 0.795, respectively (p < 0.001). For dose-volumetric parameters, MIt showed higher statistically significant correlations than did the conventional modulation indices. Conclusion: The MIt, contrast (d = 1) and variance (d = 1) showed good performance to predict the VMAT delivery accuracy showing higher correlations to the results of various types of verification methods for VMAT. This work was in part supported by the National Research Foundation of Korea (NRF) grant (no. 490-20140029 and no. 490-20130047) funded by the Korea government.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
DiCostanzo, D; Ayan, A; Woollard, J
Purpose: To predict potential failures of hardware within the Varian TrueBeam linear accelerator in order to proactively replace parts and decrease machine downtime. Methods: Machine downtime is a problem for all radiation oncology departments and vendors. Most often it is the result of unexpected equipment failure, and increased due to lack of in-house clinical engineering support. Preventative maintenance attempts to assuage downtime, but often is ineffective at preemptively preventing many failure modes such as MLC motor failures, the need to tighten a gantry chain, or the replacement of a jaw motor, among other things. To attempt to alleviate downtime, softwaremore » was developed in house that determines the maximum value of each axis enumerated in the Truebeam trajectory log files. After patient treatments, this data is stored in a SQL database. Microsoft Power BI is used to plot the average maximum error of each day of each machine as a function of time. The results are then correlated with actual faults that occurred at the machine with the help of Varian service engineers. Results: Over the course of six months, 76,312 trajectory logs have been written into the database and plotted in Power BI. Throughout the course of analysis MLC motors have been replaced on three machines due to the early warning of the trajectory log analysis. The service engineers have also been alerted to possible gantry issues on one occasion due to the aforementioned analysis. Conclusion: Analyzing the trajectory log data is a viable and effective early warning system for potential failures of the TrueBeam linear accelerator. With further analysis and tightening of the tolerance values used to determine a possible imminent failure, it should be possible to pinpoint future issues more thoroughly and for more axes of motion.« less
TU-CD-304-11: Veritas 2.0: A Cloud-Based Tool to Facilitate Research and Innovation
DOE Office of Scientific and Technical Information (OSTI.GOV)
Mishra, P; Patankar, A; Etmektzoglou, A
Purpose: We introduce Veritas 2.0, a cloud-based, non-clinical research portal, to facilitate translation of radiotherapy research ideas to new delivery techniques. The ecosystem of research tools includes web apps for a research beam builder for TrueBeam Developer Mode, an image reader for compressed and uncompressed XIM files, and a trajectory log file based QA/beam delivery analyzer. Methods: The research beam builder can generate TrueBeam readable XML file either from scratch or from pre-existing DICOM-RT plans. DICOM-RT plan is first converted to XML format and then researcher can interactively modify or add control points to them. Delivered beam can be verifiedmore » via reading generated images and analyzing trajectory log files. Image reader can read both uncompressed and HND-compressed XIM images. The trajectory log analyzer lets researchers plot expected vs. actual values and deviations among 30 mechanical axes. The analyzer gives an animated view of MLC patterns for the beam delivery. Veritas 2.0 is freely available and its advantages versus standalone software are i) No software installation or maintenance needed, ii) easy accessibility across all devices iii) seamless upgrades and iv) OS independence. Veritas is written using open-source tools like twitter bootstrap, jQuery, flask, and Python-based modules. Results: In the first experiment, an anonymized 7-beam DICOM-RT IMRT plan was converted to XML beam containing 1400 control points. kV and MV imaging points were inserted into this XML beam. In another experiment, a binary log file was analyzed to compare actual vs expected values and deviations among axes. Conclusions: Veritas 2.0 is a public cloud-based web app that hosts a pool of research tools for facilitating research from conceptualization to verification. It is aimed at providing a platform for facilitating research and collaboration. I am full time employee at Varian Medical systems, Palo Alto.« less
NASA Astrophysics Data System (ADS)
Rahman, Md Mushfiqur; Lei, Yu; Kalantzis, Georgios
2018-01-01
Quality Assurance (QA) for medical linear accelerator (linac) is one of the primary concerns in external beam radiation Therapy. Continued advancements in clinical accelerators and computer control technology make the QA procedures more complex and time consuming which often, adequate software accompanied with specific phantoms is required. To ameliorate that matter, we introduce QALMA (Quality Assurance for Linac with MATLAB), a MALAB toolkit which aims to simplify the quantitative analysis of QA for linac which includes Star-Shot analysis, Picket Fence test, Winston-Lutz test, Multileaf Collimator (MLC) log file analysis and verification of light & radiation field coincidence test.
TH-A-9A-10: Prostate SBRT Delivery with Flattening-Filter-Free Mode: Benefit and Accuracy
DOE Office of Scientific and Technical Information (OSTI.GOV)
Li, T; Yuan, L; Sheng, Y
Purpose: Flattening-filter-free (FFF) beam mode offered on TrueBeam™ linac enables delivering IMRT at 2400 MU/min dose rate. This study investigates the benefit and delivery accuracy of using high dose rate in the context of prostate SBRT. Methods: 8 prostate SBRT patients were retrospectively studied. In 5 cases treated with 600-MU/min dose rate, continuous prostate motion data acquired during radiation-beam-on was used to analyze motion range. In addition, the initial 1/3 of prostate motion trajectories during each radiation-beam-on was separated to simulate motion range if 2400-MU/min were used. To analyze delivery accuracy in FFF mode, MLC trajectory log files from anmore » additional 3 cases treated at 2400-MU/min were acquired. These log files record MLC expected and actual positions every 20ms, and therefore can be used to reveal delivery accuracy. Results: (1) Benefit. On average treatment at 600-MU/min takes 30s per beam; whereas 2400-MU/min requires only 11s. When shortening delivery time to ~1/3, the prostate motion range was significantly smaller (p<0.001). Largest motion reduction occurred in Sup-Inf direction, from [−3.3mm, 2.1mm] to [−1.7mm, 1.7mm], followed by reduction from [−2.1mm, 2.4mm] to [−1.0mm, 2.4mm] in Ant-Pos direction. No change observed in LR direction [−0.8mm, 0.6mm]. The combined motion amplitude (vector norm) confirms that average motion and ranges are significantly smaller when beam-on was limited to the 1st 1/3 of actual delivery time. (2) Accuracy. Trajectory log file analysis showed excellent delivery accuracy with at 2400 MU/min. Most leaf deviations during beam-on were within 0.07mm (99-percentile). Maximum leaf-opening deviations during each beam-on were all under 0.1mm for all leaves. Dose-rate was maintained at 2400-MU/min during beam-on without dipping. Conclusion: Delivery prostate SBRT with 2400 MU/min is both beneficial and accurate. High dose rates significantly reduced both treatment time and intra-beam prostate motion range. Excellent delivery accuracy was confirmed with very small leaf motion deviation.« less
Consistency and reproducibility of the VMAT plan delivery using three independent validation methods
Chandraraj, Varatharaj; Manickam, Ravikumar; Esquivel, Carlos; Supe, Sanjay S; Papanikolaou, Nikos
2010-01-01
The complexity of VMAT delivery requires new methods and potentially new tools for the commissioning of these systems. It appears that great consideration is needed for quality assurance (QA) of these treatments since there are limited devices that are dedicated to the QA of rotational delivery. In this present study, we have evaluated the consistency and reproducibility of one prostate and one lung VMAT plans for 31 consecutive days using three different approaches: 1) MLC DynaLog files, 2) in vivo measurements using the multiwire ionization chamber DAVID, and 3) using PTWseven29 2D ARRAY with the OCTAVIUS phantom at our Varian Clinac linear accelerator. Overall, the three methods of testing the reproducibility and consistency of the VMAT delivery were in agreement with each other. All methods showed minimal daily deviations that contributed to clinically insignificant dose variations from day to day. Based on our results, we conclude that the VMAT delivery using a Varian 2100CD linear accelerator equipped with 120 MLC is highly reproducible. PACS numbers: 87.55.Qr and 87.56.Fc
DOE Office of Scientific and Technical Information (OSTI.GOV)
Zwan, B J; University of Newcastle, Newcastle, NSW; Barnes, M
2016-06-15
Purpose: To automate gantry-resolved linear accelerator (linac) quality assurance (QA) for volumetric modulated arc therapy (VMAT) using an electronic portal imaging device (EPID). Methods: A QA system for VMAT was developed that uses an EPID, frame-grabber assembly and in-house developed image processing software. The system relies solely on the analysis of EPID image frames acquired without the presence of a phantom. Images were acquired at 8.41 frames per second using a frame grabber and ancillary acquisition computer. Each image frame was tagged with a gantry angle from the linac’s on-board gantry angle encoder. Arc-dynamic QA plans were designed to assessmore » the performance of each individual linac component during VMAT. By analysing each image frame acquired during the QA deliveries the following eight machine performance characteristics were measured as a function of gantry angle: MLC positional accuracy, MLC speed constancy, MLC acceleration constancy, MLC-gantry synchronisation, beam profile constancy, dose rate constancy, gantry speed constancy, dose-gantry angle synchronisation and mechanical sag. All tests were performed on a Varian iX linear accelerator equipped with a 120 leaf Millennium MLC and an aS1000 EPID (Varian Medical Systems, Palo Alto, CA, USA). Results: Machine performance parameters were measured as a function of gantry angle using EPID imaging and compared to machine log files and the treatment plan. Data acquisition is currently underway at 3 centres, incorporating 7 treatment units, at 2 weekly measurement intervals. Conclusion: The proposed system can be applied for streamlined linac QA and commissioning for VMAT. The set of test plans developed can be used to assess the performance of each individual components of the treatment machine during VMAT deliveries as a function of gantry angle. The methodology does not require the setup of any additional phantom or measurement equipment and the analysis is fully automated to allow for regular routine testing.« less
Park, Jong In; Park, Jong Min; Kim, Jung-In; Park, So-Yeon; Ye, Sung-Joon
2015-12-01
The aim of this study was to investigate the sensitivity of the gamma-index method according to various gamma criteria for volumetric modulated arc therapy (VMAT). Twenty head and neck (HN) and twenty prostate VMAT plans were retrospectively selected for this study. Both global and local 2D gamma evaluations were performed with criteria of 3%/3 mm, 2%/2 mm, 1%/2 mm and 2%/1 mm. In this study, the global and local gamma-index calculated the differences in doses relative to the maximum dose and the dose at the current measurement point, respectively. Using log files acquired during delivery, the differences in parameters at every control point between the VMAT plans and the log files were acquired. The differences in dose-volumetric parameters between reconstructed VMAT plans using the log files and the original VMAT plans were calculated. The Spearman's rank correlation coefficients (rs) were calculated between the passing rates and those differences. Considerable correlations with statistical significances were observed between global 1%/2 mm, local 1%/2 mm and local 2%/1 mm and the MLC position differences (rs = -0.712, -0.628 and -0.581). The numbers of rs values with statistical significance between the passing rates and the changes in dose-volumetric parameters were largest in global 2%/2 mm (n = 16), global 2%/1 mm (n = 15) and local 2%/1 mm (n = 13) criteria. Local gamma-index method with 2%/1 mm generally showed higher sensitivity to detect deviations between a VMAT plan and the delivery of the VMAT plan. Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
VMAT testing for an Elekta accelerator
Sweeney, Larry E.; Marshall, Edward I.; Mahendra, Saikanth
2012-01-01
Volumetric‐modulated arc therapy (VMAT) has been shown to be able to deliver plans equivalent to intensity‐modulated radiation therapy (IMRT) in a fraction of the treatment time. This improvement is important for patient immobilization/ localization compliance due to comfort and treatment duration, as well as patient throughput. Previous authors have suggested commissioning methods for this modality. Here, we extend the methods reported for the Varian RapidArc system (which tested individual system components) to the Elekta linear accelerator, using custom files built using the Elekta iComCAT software. We also extend the method reported for VMAT commissioning of the Elekta accelerator by verifying maximum values of parameters (gantry speed, multileaf collimator (MLC) speed, and backup jaw speed), investigating: 1) beam profiles as a function of dose rate during an arc, 2) over/under dosing due to MLC reversals, and 3) over/under dosing at changing dose rate junctions. Equations for construction of the iComCAT files are given. Results indicate that the beam profile for lower dose rates varies less than 3% from that of the maximum dose rate, with no difference during an arc. The gantry, MLC, and backup jaw maximum speed are internally consistent. The monitor unit chamber is stable over the MUs and gantry movement conditions expected. MLC movement and position during VMAT delivery are within IMRT tolerances. Dose rate, gantry speed, and MLC speed are accurately controlled. Over/under dosing at junctions of MLC reversals or dose rate changes are within clinical acceptability. PACS numbers: 87.55.de, 87.55.Qr, 87.56.bd PMID:22402389
Chow, James C.L.; Grigorov, Grigor N.; Yazdani, Nuri
2006-01-01
A custom‐made computer program, SWIMRT, to construct “multileaf collimator (MLC) machine” file for intensity‐modulated radiotherapy (IMRT) fluence maps was developed using MATLAB® and the sliding window algorithm. The user can either import a fluence map with a graphical file format created by an external treatment‐planning system such as Pinnacle3 or create his or her own fluence map using the matrix editor in the program. Through comprehensive calibrations of the dose and the dimension of the imported fluence field, the user can use associated image‐processing tools such as field resizing and edge trimming to modify the imported map. When the processed fluence map is suitable, a “MLC machine” file is generated for our Varian 21 EX linear accelerator with a 120‐leaf Millennium MLC. This machine file is transferred to the MLC console of the LINAC to control the continuous motions of the leaves during beam irradiation. An IMRT field is then irradiated with the 2D intensity profiles, and the irradiated profiles are compared to the imported or modified fluence map. This program was verified and tested using film dosimetry to address the following uncertainties: (1) the mechanical limitation due to the leaf width and maximum traveling speed, and (2) the dosimetric limitation due to the leaf leakage/transmission and penumbra effect. Because the fluence map can be edited, resized, and processed according to the requirement of a study, SWIMRT is essential in studying and investigating the IMRT technique using the sliding window algorithm. Using this program, future work on the algorithm may include redistributing the time space between segmental fields to enhance the fluence resolution, and readjusting the timing of each leaf during delivery to avoid small fields. Possible clinical utilities and examples for SWIMRT are given in this paper. PACS numbers: 87.53.Kn, 87.53.St, 87.53.Uv PMID:17533330
Ruiz-Angel, M J; Carda-Broch, S; García-Alvarez-Coque, M C; Berthod, A
2004-03-19
Logarithm of retention factors (log k) of a group of 14 ionizable diuretics were correlated with the molecular (log P o/w) and apparent (log P(app)) octanol-water partition coefficients. The compounds were chromatographed using aqueous-organic (reversed-phase liquid chromatography, RPLC) and micellar-organic mobile phases (micellar liquid chromatography, MLC) with the anionic surfactant sodium dodecyl sulfate (SDS), in the pH range 3-7, and a conventional octadecylsilane column. Acetonitrile was used as the organic modifier in both modes. The quality of the correlations obtained for log P(app) at varying ionization degree confirms that this correction is required in the aqueous-organic mixtures. The correlation is less improved with SDS micellar media because the acid-base equilibriums are shifted towards higher pH values for acidic compounds. In micellar chromatography, an electrostatic interaction with charged solutes is added to hydrophobic forces; consequently, different correlations should be established for neutral and acidic compounds, and for basic compounds. Correlations between log k and the isocratic descriptors log k(w), log k(wm) (extrapolated retention to pure water in the aqueous-organic and micellar-organic systems, respectively), and psi0 (extrapolated mobile phase composition giving a k = 1 retention factor or twice the dead time), and between these descriptors and log P(app) were also satisfactory, although poorer than those between log k and log P(app) due to the extrapolation. The study shows that, in the particular case of the ionizable diuretics studied, classical RPLC gives better results than MLC with SDS in the retention hydrophobicity correlations.
Feasibility of using the linac real-time log data for VMAT treatment verification
NASA Astrophysics Data System (ADS)
Midi, N. S.; Zin, Hafiz M.
2017-05-01
This study investigates the feasibility of using the real-time log data from a linac to verify Volumetric Modulated Arc Therapy (VMAT) treatment. The treatment log data for an Elekta Synergy linac can be recorded at a sampling rate of 4 Hz using the service graphing tool on the linac control computer. A treatment plan that simulates a VMAT treatment was delivered from the linac and all the dynamic treatment parameters including monitor unit (MU), Multileaf Collimator (MLC) position, jaw position, gantry angle and collimator angle were recorded in real-time using the service graphing tool. The recorded raw data were extracted and analysed using algorithms written in Matlab (MathWorks, Natick, MA). The actual treatment parameters logged using the service graphing tool was compared to the prescription and the deviations were analysed. The MLC position errors travelling at the speed range from -3.25 to 5.92 cm/s were between -1.7 mm to 2.5 mm, well within the 3.5 mm tolerance value (AAPM TG-142). The discrepancies of other delivery parameters were also within the tolerance. The real-time linac parameters logged using the service graphing tool can be used as a supplementary data for patient specific VMAT pre-treatment quality assurance.
Silanols, a New Class of Antimicrobial Agent
2006-04-01
carbinols against the four bacteria was log (1/MLC) = 0.670 log P + 0.0035 ∆ν -1.836, n = 282, r = 0.96, s = 0.22. This equation and a significantly...activity relationship of antimicrobial agents by means of equations [8] based on a method proposed by Hansch and Fujita in 1964 [1]. This multiple...correlation equations between their antimicrobial activities and structural properties, log P and H-bond acidity, were created by a multiple regression
DOE Office of Scientific and Technical Information (OSTI.GOV)
Ahmed, S; Kakakhel, MB; Ahmed, SBS
2015-06-15
Purpose: The primary aim was to introduce a dose optimization method for translating bed total body irradiation technique that ensures lung shielding dynamically. Symmetric and asymmetric dynamic MLC apertures were employed for this purpose. Methods: The MLC aperture sizes were defined based on the radiological depth values along the divergent ray lines passing through the individual CT slices. Based on these RD values, asymmetrically shaped MLC apertures were defined every 9 mm of the phantom in superior-inferior direction. Individual MLC files were created with MATLAB™ and were imported into Eclipse™ treatment planning system for dose calculations. Lungs can be shieldedmore » to an optimum level by reducing the MLC aperture width over the lungs. The process was repeated with symmetrically shaped apertures. Results: Dose-volume histogram (DVH) analysis shows that the asymmetric MLC based technique provides better dose coverage to the body and optimum shielding of the lungs compared to symmetrically shaped beam apertures. Midline dose homogeneity is within ±3% with asymmetric MLC apertures whereas it remains within ±4.5% with symmetric ones (except head region where it drops down to −7%). The substantial over and under dosage of ±5% at tissue interfaces has been reduced to ±2% with asymmetric MLC technique. Lungs dose can be reduced to any desired limit. In this experiment lungs dose was reduced to 80% of the prescribed dose, as was desired. Conclusion: The novel asymmetric MLC based technique assures optimum shielding of OARs (e.g. lungs) and better 3-D dose homogeneity and body-dose coverage in comparison with the symmetric MLC aperture optimization. The authors acknowledge the financial and infrastructural support provided by Pakistan Institute of Engineering & Applied Sciences (PIEAS), Islamabad and Aga Khan University Hospital (AKUH), Karachi during the course of this research project. Authors have no conflict of interest with any national / international body for the presented work.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Gardner, S; Gulam, M; Song, K
2014-06-01
Purpose: The Varian EDGE machine is a new stereotactic platform, combining Calypso and VisionRT localization systems with a stereotactic linac. The system includes TrueBeam DeveloperMode, making possible the use of XML-scripting for automation of linac-related tasks. This study details the use of DeveloperMode to automate commissioning tasks for Varian EDGE, thereby improving efficiency and measurement consistency. Methods: XML-scripting was used for various commissioning tasks,including couch model verification,beam-scanning,and isocenter verification. For couch measurements, point measurements were acquired for several field sizes (2×2,4×4,10×10cm{sup 2}) at 42 gantry angles for two couch-models. Measurements were acquired with variations in couch position(rails in/out,couch shifted inmore » each of motion axes) compared to treatment planning system(TPS)-calculated values,which were logged automatically through advanced planning interface(API) scripting functionality. For beam scanning, XML-scripts were used to create custom MLC-apertures. For isocenter verification, XML-scripts were used to automate various Winston-Lutz-type tests. Results: For couch measurements, the time required for each set of angles was approximately 9 minutes. Without scripting, each set required approximately 12 minutes. Automated measurements required only one physicist, while manual measurements required at least two physicists to handle linac positions/beams and data recording. MLC apertures were generated outside of the TPS,and with the .xml file format, double-checking without use of TPS/operator console was possible. Similar time efficiency gains were found for isocenter verification measurements Conclusion: The use of XML scripting in TrueBeam DeveloperMode allows for efficient and accurate data acquisition during commissioning. The efficiency improvement is most pronounced for iterative measurements, exemplified by the time savings for couch modeling measurements(approximately 10 hours). The scripting also allowed for creation of the files in advance without requiring access to TPS. The API scripting functionality enabled efficient creation/mining of TPS data. Finally, automation reduces the potential for human error in entering linac values at the machine console,and the script provides a log of measurements acquired for each session. This research was supported in part by a grant from Varian Medical Systems, Palo Alto, CA.« less
Who Goes There? Measuring Library Web Site Usage.
ERIC Educational Resources Information Center
Bauer, Kathleen
2000-01-01
Discusses how libraries can gather data on the use of their Web sites. Highlights include Web server log files, including the common log file, referrer log file, and agent log file; log file limitations; privacy concerns; and choosing log analysis software, both free and commercial. (LRW)
Wooten, H. Omar; Green, Olga; Li, Harold H.; Liu, Shi; Li, Xiaoling; Rodriguez, Vivian; Mutic, Sasa; Kashani, Rojano
2016-01-01
The aims of this study were to develop a method for automatic and immediate verification of treatment delivery after each treatment fraction in order to detect and correct errors, and to develop a comprehensive daily report which includes delivery verification results, daily image‐guided radiation therapy (IGRT) review, and information for weekly physics reviews. After systematically analyzing the requirements for treatment delivery verification and understanding the available information from a commercial MRI‐guided radiotherapy treatment machine, we designed a procedure to use 1) treatment plan files, 2) delivery log files, and 3) beam output information to verify the accuracy and completeness of each daily treatment delivery. The procedure verifies the correctness of delivered treatment plan parameters including beams, beam segments and, for each segment, the beam‐on time and MLC leaf positions. For each beam, composite primary fluence maps are calculated from the MLC leaf positions and segment beam‐on time. Error statistics are calculated on the fluence difference maps between the plan and the delivery. A daily treatment delivery report is designed to include all required information for IGRT and weekly physics reviews including the plan and treatment fraction information, daily beam output information, and the treatment delivery verification results. A computer program was developed to implement the proposed procedure of the automatic delivery verification and daily report generation for an MRI guided radiation therapy system. The program was clinically commissioned. Sensitivity was measured with simulated errors. The final version has been integrated into the commercial version of the treatment delivery system. The method automatically verifies the EBRT treatment deliveries and generates the daily treatment reports. Already in clinical use for over one year, it is useful to facilitate delivery error detection, and to expedite physician daily IGRT review and physicist weekly chart review. PACS number(s): 87.55.km PMID:27167269
75 FR 78179 - Telemarketing Sales Rule
Federal Register 2010, 2011, 2012, 2013, 2014
2010-12-15
... Venkataraman, Civ. No. 3:06-cv-01928-MLC- JJH (D. N.J. filed Apr. 26, 2006); United States v. Civic Development... that initiated the telephone call. \\39\\ Complaint ] 17, United States v. Srikanth Venkataraman (d/ b/a...
De Vrieze, Mike; Lynen, Frédéric; Chen, Kai; Szucs, Roman; Sandra, Pat
2013-07-01
Several in vitro methods have been tested for their ability to predict drug penetration across the blood-brain barrier (BBB) into the central nervous system (CNS). In this article, the performance of a variety of micellar liquid chromatographic (MLC) methods and immobilized artificial membrane (IAM) liquid chromatographic approaches were compared for a set of 45 solutes. MLC measurements were performed on a C18 column with sodium dodecyl sulfate (SDS), polyoxyethylene (23) lauryl ether (Brij35), or sodium deoxycholate (SDC) as surfactant in the micellar mobile phase. IAM liquid chromatography measurements were performed with Dulbecco's phosphate-buffered saline (DPBS) and methanol as organic modifier in the mobile phase. The corresponding retention and computed descriptor data for each solute were used for construction of models to predict transport across the blood-brain barrier (log BB). All data were correlated with experimental log BB values and the relative performance of the models was studied. SDS-based models proved most suitable for prediction of log BB values, followed closely by a simplified IAM method, in which it could be observed that extrapolation of retention data to 0% modifier in the mobile phase was unnecessary.
Software tool for portal dosimetry research.
Vial, P; Hunt, P; Greer, P B; Oliver, L; Baldock, C
2008-09-01
This paper describes a software tool developed for research into the use of an electronic portal imaging device (EPID) to verify dose for intensity modulated radiation therapy (IMRT) beams. A portal dose image prediction (PDIP) model that predicts the EPID response to IMRT beams has been implemented into a commercially available treatment planning system (TPS). The software tool described in this work was developed to modify the TPS PDIP model by incorporating correction factors into the predicted EPID image to account for the difference in EPID response to open beam radiation and multileaf collimator (MLC) transmitted radiation. The processes performed by the software tool include; i) read the MLC file and the PDIP from the TPS, ii) calculate the fraction of beam-on time that each point in the IMRT beam is shielded by MLC leaves, iii) interpolate correction factors from look-up tables, iv) create a corrected PDIP image from the product of the original PDIP and the correction factors and write the corrected image to file, v) display, analyse, and export various image datasets. The software tool was developed using the Microsoft Visual Studio.NET framework with the C# compiler. The operation of the software tool was validated. This software provided useful tools for EPID dosimetry research, and it is being utilised and further developed in ongoing EPID dosimetry and IMRT dosimetry projects.
The sensitivity of patient specific IMRT QC to systematic MLC leaf bank offset errors.
Rangel, Alejandra; Palte, Gesa; Dunscombe, Peter
2010-07-01
Patient specific IMRT QC is performed routinely in many clinics as a safeguard against errors and inaccuracies which may be introduced during the complex planning, data transfer, and delivery phases of this type of treatment. The purpose of this work is to evaluate the feasibility of detecting systematic errors in MLC leaf bank position with patient specific checks. 9 head and neck (H&N) and 14 prostate IMRT beams were delivered using MLC files containing systematic offsets (+/- 1 mm in two banks, +/- 0.5 mm in two banks, and 1 mm in one bank of leaves). The beams were measured using both MAPCHECK (Sun Nuclear Corp., Melbourne, FL) and the aS1000 electronic portal imaging device (Varian Medical Systems, Palo Alto, CA). Comparisons with calculated fields, without offsets, were made using commonly adopted criteria including absolute dose (AD) difference, relative dose difference, distance to agreement (DTA), and the gamma index. The criteria most sensitive to systematic leaf bank offsets were the 3% AD, 3 mm DTA for MAPCHECK and the gamma index with 2% AD and 2 mm DTA for the EPID. The criterion based on the relative dose measurements was the least sensitive to MLC offsets. More highly modulated fields, i.e., H&N, showed greater changes in the percentage of passing points due to systematic MLC inaccuracy than prostate fields. None of the techniques or criteria tested is sufficiently sensitive, with the population of IMRT fields, to detect a systematic MLC offset at a clinically significant level on an individual field. Patient specific QC cannot, therefore, substitute for routine QC of the MLC itself.
Determination of the optimal tolerance for MLC positioning in sliding window and VMAT techniques
DOE Office of Scientific and Technical Information (OSTI.GOV)
Hernandez, V., E-mail: vhernandezmasgrau@gmail.com; Abella, R.; Calvo, J. F.
2015-04-15
Purpose: Several authors have recommended a 2 mm tolerance for multileaf collimator (MLC) positioning in sliding window treatments. In volumetric modulated arc therapy (VMAT) treatments, however, the optimal tolerance for MLC positioning remains unknown. In this paper, the authors present the results of a multicenter study to determine the optimal tolerance for both techniques. Methods: The procedure used is based on dynalog file analysis. The study was carried out using seven Varian linear accelerators from five different centers. Dynalogs were collected from over 100 000 clinical treatments and in-house software was used to compute the number of tolerance faults as amore » function of the user-defined tolerance. Thus, the optimal value for this tolerance, defined as the lowest achievable value, was investigated. Results: Dynalog files accurately predict the number of tolerance faults as a function of the tolerance value, especially for low fault incidences. All MLCs behaved similarly and the Millennium120 and the HD120 models yielded comparable results. In sliding window techniques, the number of beams with an incidence of hold-offs >1% rapidly decreases for a tolerance of 1.5 mm. In VMAT techniques, the number of tolerance faults sharply drops for tolerances around 2 mm. For a tolerance of 2.5 mm, less than 0.1% of the VMAT arcs presented tolerance faults. Conclusions: Dynalog analysis provides a feasible method for investigating the optimal tolerance for MLC positioning in dynamic fields. In sliding window treatments, the tolerance of 2 mm was found to be adequate, although it can be reduced to 1.5 mm. In VMAT treatments, the typically used 5 mm tolerance is excessively high. Instead, a tolerance of 2.5 mm is recommended.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Wisotzky, Eric, E-mail: eric.wisotzky@charite.de, E-mail: eric.wisotzky@ipk.fraunhofer.de; O’Brien, Ricky; Keall, Paul J., E-mail: paul.keall@sydney.edu.au
2016-01-15
Purpose: Multileaf collimator (MLC) tracking radiotherapy is complex as the beam pattern needs to be modified due to the planned intensity modulation as well as the real-time target motion. The target motion cannot be planned; therefore, the modified beam pattern differs from the original plan and the MLC sequence needs to be recomputed online. Current MLC tracking algorithms use a greedy heuristic in that they optimize for a given time, but ignore past errors. To overcome this problem, the authors have developed and improved an algorithm that minimizes large underdose and overdose regions. Additionally, previous underdose and overdose events aremore » taken into account to avoid regions with high quantity of dose events. Methods: The authors improved the existing MLC motion control algorithm by introducing a cumulative underdose/overdose map. This map represents the actual projection of the planned tumor shape and logs occurring dose events at each specific regions. These events have an impact on the dose cost calculation and reduce recurrence of dose events at each region. The authors studied the improvement of the new temporal optimization algorithm in terms of the L1-norm minimization of the sum of overdose and underdose compared to not accounting for previous dose events. For evaluation, the authors simulated the delivery of 5 conformal and 14 intensity-modulated radiotherapy (IMRT)-plans with 7 3D patient measured tumor motion traces. Results: Simulations with conformal shapes showed an improvement of L1-norm up to 8.5% after 100 MLC modification steps. Experiments showed comparable improvements with the same type of treatment plans. Conclusions: A novel leaf sequencing optimization algorithm which considers previous dose events for MLC tracking radiotherapy has been developed and investigated. Reductions in underdose/overdose are observed for conformal and IMRT delivery.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Li, Taoran, E-mail: taoran.li.duke@gmail.com; Wu, Qiuwen; Yang, Yun
Purpose: An important challenge facing online adaptive radiation therapy is the development of feasible and efficient quality assurance (QA). This project aimed to validate the deliverability of online adapted plans and develop a proof-of-concept online delivery monitoring system for online adaptive radiation therapy QA. Methods: The first part of this project benchmarked automatically online adapted prostate treatment plans using traditional portal dosimetry IMRT QA. The portal dosimetry QA results of online adapted plans were compared to original (unadapted) plans as well as randomly selected prostate IMRT plans from our clinic. In the second part, an online delivery monitoring system wasmore » designed and validated via a simulated treatment with intentional multileaf collimator (MLC) errors. This system was based on inputs from the dynamic machine information (DMI), which continuously reports actual MLC positions and machine monitor units (MUs) at intervals of 50 ms or less during delivery. Based on the DMI, the system performed two levels of monitoring/verification during the delivery: (1) dynamic monitoring of cumulative fluence errors resulting from leaf position deviations and visualization using fluence error maps (FEMs); and (2) verification of MLC positions against the treatment plan for potential errors in MLC motion and data transfer at each control point. Validation of the online delivery monitoring system was performed by introducing intentional systematic MLC errors (ranging from 0.5 to 2 mm) to the DMI files for both leaf banks. These DMI files were analyzed by the proposed system to evaluate the system’s performance in quantifying errors and revealing the source of errors, as well as to understand patterns in the FEMs. In addition, FEMs from 210 actual prostate IMRT beams were analyzed using the proposed system to further validate its ability to catch and identify errors, as well as establish error magnitude baselines for prostate IMRT delivery. Results: Online adapted plans were found to have similar delivery accuracy in comparison to clinical IMRT plans when validated with portal dosimetry IMRT QA. FEMs for the simulated deliveries with intentional MLC errors exhibited distinct patterns for different MLC error magnitudes and directions, indicating that the proposed delivery monitoring system is highly specific in detecting the source of errors. Implementing the proposed QA system for online adapted plans revealed excellent delivery accuracy: over 99% of leaf position differences were within 0.5 mm, and >99% of pixels in the FEMs had fluence errors within 0.5 MU. Patterns present in the FEMs and MLC control point analysis for actual patient cases agreed with the error pattern analysis results, further validating the system’s ability to reveal and differentiate MLC deviations. Calculation of the fluence map based on the DMI was performed within 2 ms after receiving each DMI input. Conclusions: The proposed online delivery monitoring system requires minimal additional resources and time commitment to the current clinical workflow while still maintaining high sensitivity to leaf position errors and specificity to error types. The presented online delivery monitoring system therefore represents a promising QA system candidate for online adaptive radiation therapy.« less
Li, Taoran; Wu, Qiuwen; Yang, Yun; Rodrigues, Anna; Yin, Fang-Fang; Jackie Wu, Q
2015-01-01
An important challenge facing online adaptive radiation therapy is the development of feasible and efficient quality assurance (QA). This project aimed to validate the deliverability of online adapted plans and develop a proof-of-concept online delivery monitoring system for online adaptive radiation therapy QA. The first part of this project benchmarked automatically online adapted prostate treatment plans using traditional portal dosimetry IMRT QA. The portal dosimetry QA results of online adapted plans were compared to original (unadapted) plans as well as randomly selected prostate IMRT plans from our clinic. In the second part, an online delivery monitoring system was designed and validated via a simulated treatment with intentional multileaf collimator (MLC) errors. This system was based on inputs from the dynamic machine information (DMI), which continuously reports actual MLC positions and machine monitor units (MUs) at intervals of 50 ms or less during delivery. Based on the DMI, the system performed two levels of monitoring/verification during the delivery: (1) dynamic monitoring of cumulative fluence errors resulting from leaf position deviations and visualization using fluence error maps (FEMs); and (2) verification of MLC positions against the treatment plan for potential errors in MLC motion and data transfer at each control point. Validation of the online delivery monitoring system was performed by introducing intentional systematic MLC errors (ranging from 0.5 to 2 mm) to the DMI files for both leaf banks. These DMI files were analyzed by the proposed system to evaluate the system's performance in quantifying errors and revealing the source of errors, as well as to understand patterns in the FEMs. In addition, FEMs from 210 actual prostate IMRT beams were analyzed using the proposed system to further validate its ability to catch and identify errors, as well as establish error magnitude baselines for prostate IMRT delivery. Online adapted plans were found to have similar delivery accuracy in comparison to clinical IMRT plans when validated with portal dosimetry IMRT QA. FEMs for the simulated deliveries with intentional MLC errors exhibited distinct patterns for different MLC error magnitudes and directions, indicating that the proposed delivery monitoring system is highly specific in detecting the source of errors. Implementing the proposed QA system for online adapted plans revealed excellent delivery accuracy: over 99% of leaf position differences were within 0.5 mm, and >99% of pixels in the FEMs had fluence errors within 0.5 MU. Patterns present in the FEMs and MLC control point analysis for actual patient cases agreed with the error pattern analysis results, further validating the system's ability to reveal and differentiate MLC deviations. Calculation of the fluence map based on the DMI was performed within 2 ms after receiving each DMI input. The proposed online delivery monitoring system requires minimal additional resources and time commitment to the current clinical workflow while still maintaining high sensitivity to leaf position errors and specificity to error types. The presented online delivery monitoring system therefore represents a promising QA system candidate for online adaptive radiation therapy.
TH-AB-201-12: Using Machine Log-Files for Treatment Planning and Delivery QA
DOE Office of Scientific and Technical Information (OSTI.GOV)
Stanhope, C; Liang, J; Drake, D
2016-06-15
Purpose: To determine the segment reduction and dose resolution necessary for machine log-files to effectively replace current phantom-based patient-specific quality assurance, while minimizing computational cost. Methods: Elekta’s Log File Convertor R3.2 records linac delivery parameters (dose rate, gantry angle, leaf position) every 40ms. Five VMAT plans [4 H&N, 1 Pulsed Brain] comprised of 2 arcs each were delivered on the ArcCHECK phantom. Log-files were reconstructed in Pinnacle on the phantom geometry using 1/2/3/4° control point spacing and 2/3/4mm dose grid resolution. Reconstruction effectiveness was quantified by comparing 2%/2mm gamma passing rates of the original and log-file plans. Modulation complexity scoresmore » (MCS) were calculated for each beam to correlate reconstruction accuracy and beam modulation. Percent error in absolute dose for each plan-pair combination (log-file vs. ArcCHECK, original vs. ArcCHECK, log-file vs. original) was calculated for each arc and every diode greater than 10% of the maximum measured dose (per beam). Comparing standard deviations of the three plan-pair distributions, relative noise of the ArcCHECK and log-file systems was elucidated. Results: The original plans exhibit a mean passing rate of 95.1±1.3%. The eight more modulated H&N arcs [MCS=0.088±0.014] and two less modulated brain arcs [MCS=0.291±0.004] yielded log-file pass rates most similar to the original plan when using 1°/2mm [0.05%±1.3% lower] and 2°/3mm [0.35±0.64% higher] log-file reconstructions respectively. Log-file and original plans displayed percent diode dose errors 4.29±6.27% and 3.61±6.57% higher than measurement. Excluding the phantom eliminates diode miscalibration and setup errors; log-file dose errors were 0.72±3.06% higher than the original plans – significantly less noisy. Conclusion: For log-file reconstructed VMAT arcs, 1° control point spacing and 2mm dose resolution is recommended, however, less modulated arcs may allow less stringent reconstructions. Following the aforementioned reconstruction recommendations, the log-file technique is capable of detecting delivery errors with equivalent accuracy and less noise than ArcCHECK QA. I am funded by an Elekta Research Grant.« less
Quantitative evaluation of patient-specific quality assurance using online dosimetry system
NASA Astrophysics Data System (ADS)
Jung, Jae-Yong; Shin, Young-Ju; Sohn, Seung-Chang; Min, Jung-Whan; Kim, Yon-Lae; Kim, Dong-Su; Choe, Bo-Young; Suh, Tae-Suk
2018-01-01
In this study, we investigated the clinical performance of an online dosimetry system (Mobius FX system, MFX) by 1) dosimetric plan verification using gamma passing rates and dose volume metrics and 2) error-detection capability evaluation by deliberately introduced machine error. Eighteen volumetric modulated arc therapy (VMAT) plans were studied. To evaluate the clinical performance of the MFX, we used gamma analysis and dose volume histogram (DVH) analysis. In addition, to evaluate the error-detection capability, we used gamma analysis and DVH analysis utilizing three types of deliberately introduced errors (Type 1: gantry angle-independent multi-leaf collimator (MLC) error, Type 2: gantry angle-dependent MLC error, and Type 3: gantry angle error). A dosimetric verification comparison of physical dosimetry system (Delt4PT) and online dosimetry system (MFX), gamma passing rates of the two dosimetry systems showed very good agreement with treatment planning system (TPS) calculation. For the average dose difference between the TPS calculation and the MFX measurement, most of the dose metrics showed good agreement within a tolerance of 3%. For the error-detection comparison of Delta4PT and MFX, the gamma passing rates of the two dosimetry systems did not meet the 90% acceptance criterion with the magnitude of error exceeding 2 mm and 1.5 ◦, respectively, for error plans of Types 1, 2, and 3. For delivery with all error types, the average dose difference of PTV due to error magnitude showed good agreement between calculated TPS and measured MFX within 1%. Overall, the results of the online dosimetry system showed very good agreement with those of the physical dosimetry system. Our results suggest that a log file-based online dosimetry system is a very suitable verification tool for accurate and efficient clinical routines for patient-specific quality assurance (QA).
Log file-based patient dose calculations of double-arc VMAT for head-and-neck radiotherapy.
Katsuta, Yoshiyuki; Kadoya, Noriyuki; Fujita, Yukio; Shimizu, Eiji; Majima, Kazuhiro; Matsushita, Haruo; Takeda, Ken; Jingu, Keiichi
2018-04-01
The log file-based method cannot display dosimetric changes due to linac component miscalibration because of the insensitivity of log files to linac component miscalibration. The purpose of this study was to supply dosimetric changes in log file-based patient dose calculations for double-arc volumetric-modulated arc therapy (VMAT) in head-and-neck cases. Fifteen head-and-neck cases participated in this study. For each case, treatment planning system (TPS) doses were produced by double-arc and single-arc VMAT. Miscalibration-simulated log files were generated by inducing a leaf miscalibration of ±0.5 mm into the log files that were acquired during VMAT irradiation. Subsequently, patient doses were estimated using the miscalibration-simulated log files. For double-arc VMAT, regarding planning target volume (PTV), the change from TPS dose to miscalibration-simulated log file dose in D mean was 0.9 Gy and that for tumor control probability was 1.4%. As for organ-at-risks (OARs), the change in D mean was <0.7 Gy and normal tissue complication probability was <1.8%. A comparison between double-arc and single-arc VMAT for PTV showed statistically significant differences in the changes evaluated by D mean and radiobiological metrics (P < 0.01), even though the magnitude of these differences was small. Similarly, for OARs, the magnitude of these changes was found to be small. Using the log file-based method for PTV and OARs, the log file-based method estimate of patient dose using the double-arc VMAT has accuracy comparable to that obtained using the single-arc VMAT. Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
A novel method for routine quality assurance of volumetric-modulated arc therapy.
Wang, Qingxin; Dai, Jianrong; Zhang, Ke
2013-10-01
Volumetric-modulated arc therapy (VMAT) is delivered through synchronized variation of gantry angle, dose rate, and multileaf collimator (MLC) leaf positions. The delivery dynamic nature challenges the parameter setting accuracy of linac control system. The purpose of this study was to develop a novel method for routine quality assurance (QA) of VMAT linacs. ArcCheck is a detector array with diodes distributing in spiral pattern on cylindrical surface. Utilizing its features, a QA plan was designed to strictly test all varying parameters during VMAT delivery on an Elekta Synergy linac. In this plan, there are 24 control points. The gantry rotates clockwise from 181° to 179°. The dose rate, gantry speed, and MLC positions cover their ranges commonly used in clinic. The two borders of MLC-shaped field seat over two columns of diodes of ArcCheck when the gantry rotates to the angle specified by each control point. The ratio of dose rate between each of these diodes and the diode closest to the field center is a certain value and sensitive to the MLC positioning error of the leaf crossing the diode. Consequently, the positioning error can be determined by the ratio with the help of a relationship curve. The time when the gantry reaches the angle specified by each control point can be acquired from the virtual inclinometer that is a feature of ArcCheck. The gantry speed between two consecutive control points is then calculated. The aforementioned dose rate is calculated from an acm file that is generated during ArcCheck measurements. This file stores the data measured by each detector in 50 ms updates with each update in a separate row. A computer program was written in MATLAB language to process the data. The program output included MLC positioning errors and the dose rate at each control point as well as the gantry speed between control points. To evaluate this method, this plan was delivered for four consecutive weeks. The actual dose rate and gantry speed were compared with the QA plan specified. Additionally, leaf positioning errors were intentionally introduced to investigate the sensitivity of this method. The relationship curves were established for detecting MLC positioning errors during VMAT delivery. For four consecutive weeks measured, 98.4%, 94.9%, 89.2%, and 91.0% of the leaf positioning errors were within ± 0.5 mm, respectively. For the intentionally introduced leaf positioning systematic errors of -0.5 and +1 mm, the detected leaf positioning errors of 20 Y1 leaf were -0.48 ± 0.14 and 1.02 ± 0.26 mm, respectively. The actual gantry speed and dose rate closely followed the values specified in the VMAT QA plan. This method can assess the accuracy of MLC positions and the dose rate at each control point as well as the gantry speed between control points at the same time. It is efficient and suitable for routine quality assurance of VMAT.
Katsuta, Yoshiyuki; Kadoya, Noriyuki; Fujita, Yukio; Shimizu, Eiji; Matsunaga, Kenichi; Matsushita, Haruo; Majima, Kazuhiro; Jingu, Keiichi
2017-10-01
A log file-based method cannot detect dosimetric changes due to linac component miscalibration because log files are insensitive to miscalibration. Herein, clinical impacts of dosimetric changes on a log file-based method were determined. Five head-and-neck and five prostate plans were applied. Miscalibration-simulated log files were generated by inducing a linac component miscalibration into the log file. Miscalibration magnitudes for leaf, gantry, and collimator at the general tolerance level were ±0.5mm, ±1°, and ±1°, respectively, and at a tighter tolerance level achievable on current linac were ±0.3mm, ±0.5°, and ±0.5°, respectively. Re-calculations were performed on patient anatomy using log file data. Changes in tumor control probability/normal tissue complication probability from treatment planning system dose to re-calculated dose at the general tolerance level was 1.8% on planning target volume (PTV) and 2.4% on organs at risk (OARs) in both plans. These changes at the tighter tolerance level were improved to 1.0% on PTV and to 1.5% on OARs, with a statistically significant difference. We determined the clinical impacts of dosimetric changes on a log file-based method using a general tolerance level and a tighter tolerance level for linac miscalibration and found that a tighter tolerance level significantly improved the accuracy of the log file-based method. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
Bent, John M.; Faibish, Sorin; Grider, Gary
2016-04-19
Cloud object storage is enabled for checkpoints of high performance computing applications using a middleware process. A plurality of files, such as checkpoint files, generated by a plurality of processes in a parallel computing system are stored by obtaining said plurality of files from said parallel computing system; converting said plurality of files to objects using a log structured file system middleware process; and providing said objects for storage in a cloud object storage system. The plurality of processes may run, for example, on a plurality of compute nodes. The log structured file system middleware process may be embodied, for example, as a Parallel Log-Structured File System (PLFS). The log structured file system middleware process optionally executes on a burst buffer node.
SU-G-BRB-02: An Open-Source Software Analysis Library for Linear Accelerator Quality Assurance
DOE Office of Scientific and Technical Information (OSTI.GOV)
Kerns, J; Yaldo, D
Purpose: Routine linac quality assurance (QA) tests have become complex enough to require automation of most test analyses. A new data analysis software library was built that allows physicists to automate routine linear accelerator quality assurance tests. The package is open source, code tested, and benchmarked. Methods: Images and data were generated on a TrueBeam linac for the following routine QA tests: VMAT, starshot, CBCT, machine logs, Winston Lutz, and picket fence. The analysis library was built using the general programming language Python. Each test was analyzed with the library algorithms and compared to manual measurements taken at the timemore » of acquisition. Results: VMAT QA results agreed within 0.1% between the library and manual measurements. Machine logs (dynalogs & trajectory logs) were successfully parsed; mechanical axis positions were verified for accuracy and MLC fluence agreed well with EPID measurements. CBCT QA measurements were within 10 HU and 0.2mm where applicable. Winston Lutz isocenter size measurements were within 0.2mm of TrueBeam’s Machine Performance Check. Starshot analysis was within 0.2mm of the Winston Lutz results for the same conditions. Picket fence images with and without a known error showed that the library was capable of detecting MLC offsets within 0.02mm. Conclusion: A new routine QA software library has been benchmarked and is available for use by the community. The library is open-source and extensible for use in larger systems.« less
Comparing Web and Touch Screen Transaction Log Files
Huntington, Paul; Williams, Peter
2001-01-01
Background Digital health information is available on a wide variety of platforms including PC-access of the Internet, Wireless Application Protocol phones, CD-ROMs, and touch screen public kiosks. All these platforms record details of user sessions in transaction log files, and there is a growing body of research into the evaluation of this data. However, there is very little research that has examined the problems of comparing the transaction log files of kiosks and the Internet. Objectives To provide a first step towards examining the problems of comparing the transaction log files of kiosks and the Internet. Methods We studied two platforms: touch screen kiosks and a comparable Web site. For both of these platforms, we examined the menu structure (which affects transaction log file data), the log-file structure, and the metrics derived from log-file records. Results We found substantial differences between the generated metrics. Conclusions None of the metrics discussed can be regarded as an effective way of comparing the use of kiosks and Web sites. Two metrics stand out as potentially comparable and valuable: the number of user sessions per hour and user penetration of pages. PMID:11720960
DOE Office of Scientific and Technical Information (OSTI.GOV)
Katsuta, Y; Kadoya, N; Shimizu, E
2015-06-15
Purpose: A successful VMAT plan delivery includes precise modulations of dose rate, gantry rotational and multi-leaf collimator shapes. The purpose of this research is to construct routine QA protocol which focuses on VMAT delivery technique and to obtain a baseline including dose error, fluence distribution and mechanical accuracy during VMAT. Methods: The mock prostate, head and neck (HN) cases supplied from AAPM were used in this study. A VMAT plans were generated in Monaco TPS according to TG-119 protocol. Plans were created using 6 MV and 10 MV photon beams for each case. The phantom based measurement, fluence measurement andmore » log files analysis were performed. The dose measurement was performed using 0.6 cc ion chamber, which located at isocenter. The fluence distribution were acquired using the MapCHECK2 mounted in the MapPHAN. The trajectory log files recorded inner 20 leaf pairs and gantry angle positions at every 0.25 sec interval were exported to in-house software developed by MATLAB and determined those RMS values. Results: The dose difference is expressed as a ratio of the difference between measured and planned doses. The dose difference for 6 MV was 0.91%, for 10 MV was 0.67%. In turn, the fluence distribution using gamma criteria of 2%/2 mm with a 50% minimum dose threshold for 6 MV was 98.8%, for 10 MV was 97.5%, respectively. The RMS values of MLC for 6 MV and 10 MV were 0.32 mm and 0.37 mm, of gantry were 0.33 degree and 0.31 degree. Conclusion: In this study, QA protocol to assess VMAT delivery accuracy is constructed and results acquired in this study are used as a baseline of VMAT delivery performance verification.« less
Bent, John M.; Faibish, Sorin; Grider, Gary
2015-06-30
Cloud object storage is enabled for archived data, such as checkpoints and results, of high performance computing applications using a middleware process. A plurality of archived files, such as checkpoint files and results, generated by a plurality of processes in a parallel computing system are stored by obtaining the plurality of archived files from the parallel computing system; converting the plurality of archived files to objects using a log structured file system middleware process; and providing the objects for storage in a cloud object storage system. The plurality of processes may run, for example, on a plurality of compute nodes. The log structured file system middleware process may be embodied, for example, as a Parallel Log-Structured File System (PLFS). The log structured file system middleware process optionally executes on a burst buffer node.
ERIC Educational Resources Information Center
Hao, Jiangang; Smith, Lawrence; Mislevy, Robert; von Davier, Alina; Bauer, Malcolm
2016-01-01
Extracting information efficiently from game/simulation-based assessment (G/SBA) logs requires two things: a well-structured log file and a set of analysis methods. In this report, we propose a generic data model specified as an extensible markup language (XML) schema for the log files of G/SBAs. We also propose a set of analysis methods for…
Teaching an Old Log New Tricks with Machine Learning.
Schnell, Krista; Puri, Colin; Mahler, Paul; Dukatz, Carl
2014-03-01
To most people, the log file would not be considered an exciting area in technology today. However, these relatively benign, slowly growing data sources can drive large business transformations when combined with modern-day analytics. Accenture Technology Labs has built a new framework that helps to expand existing vendor solutions to create new methods of gaining insights from these benevolent information springs. This framework provides a systematic and effective machine-learning mechanism to understand, analyze, and visualize heterogeneous log files. These techniques enable an automated approach to analyzing log content in real time, learning relevant behaviors, and creating actionable insights applicable in traditionally reactive situations. Using this approach, companies can now tap into a wealth of knowledge residing in log file data that is currently being collected but underutilized because of its overwhelming variety and volume. By using log files as an important data input into the larger enterprise data supply chain, businesses have the opportunity to enhance their current operational log management solution and generate entirely new business insights-no longer limited to the realm of reactive IT management, but extending from proactive product improvement to defense from attacks. As we will discuss, this solution has immediate relevance in the telecommunications and security industries. However, the most forward-looking companies can take it even further. How? By thinking beyond the log file and applying the same machine-learning framework to other log file use cases (including logistics, social media, and consumer behavior) and any other transactional data source.
WE-G-213CD-03: A Dual Complementary Verification Method for Dynamic Tumor Tracking on Vero SBRT.
Poels, K; Depuydt, T; Verellen, D; De Ridder, M
2012-06-01
to use complementary cine EPID and gimbals log file analysis for in-vivo tracking accuracy monitoring. A clinical prototype of dynamic tracking (DT) was installed on the Vero SBRT system. This prototype version allowed tumor tracking by gimballed linac rotations using an internal-external correspondence model. The DT prototype software allowed the detailed logging of all applied gimbals rotations during tracking. The integration of an EPID on the vero system allowed the acquisition of cine EPID images during DT. We quantified the tracking error on cine EPID (E-EPID) by subtracting the target center (fiducial marker detection) and the field centroid. Dynamic gimbals log file information was combined with orthogonal x-ray verification images to calculate the in-vivo tracking error (E-kVLog). The correlation between E-kVLog and E-EPID was calculated for validation of the gimbals log file. Further, we investigated the sensitivity of the log file tracking error by introducing predefined systematic tracking errors. As an application we calculate gimbals log file tracking error for dynamic hidden target tests to investigate gravity effects and decoupled gimbals rotation from gantry rotation. Finally, calculating complementary cine EPID and log file tracking errors evaluated the clinical accuracy of dynamic tracking. A strong correlation was found between log file and cine EPID tracking error distribution during concurrent measurements (R=0.98). We found sensitivity in the gimbals log files to detect a systematic tracking error up to 0.5 mm. Dynamic hidden target tests showed no gravity influence on tracking performance and high degree of decoupled gimbals and gantry rotation during dynamic arc dynamic tracking. A submillimetric agreement between clinical complementary tracking error measurements was found. Redundancy of the internal gimbals log file with x-ray verification images with complementary independent cine EPID images was implemented to monitor the accuracy of gimballed tumor tracking on Vero SBRT. Research was financially supported by the Flemish government (FWO), Hercules Foundation and BrainLAB AG. © 2012 American Association of Physicists in Medicine.
ERIC Educational Resources Information Center
Lee, Young-Jin
2015-01-01
This study investigates whether information saved in the log files of a computer-based tutor can be used to predict the problem solving performance of students. The log files of a computer-based physics tutoring environment called Andes Physics Tutor was analyzed to build a logistic regression model that predicted success and failure of students'…
Gete, Ermias; Duzenli, Cheryl; Teke, Tony
2014-01-01
A Monte Carlo (MC) validation of the vendor‐supplied Varian TrueBeam 6 MV flattened (6X) phase‐space file and the first implementation of the Siebers‐Keall MC MLC model as applied to the HD120 MLC (for 6X flat and 6X flattening filterfree (6X FFF) beams) are described. The MC model is validated in the context of VMAT patient‐specific quality assurance. The Monte Carlo commissioning process involves: 1) validating the calculated open‐field percentage depth doses (PDDs), profiles, and output factors (OF), 2) adapting the Siebers‐Keall MLC model to match the new HD120‐MLC geometry and material composition, 3) determining the absolute dose conversion factor for the MC calculation, and 4) validating this entire linac/MLC in the context of dose calculation verification for clinical VMAT plans. MC PDDs for the 6X beams agree with the measured data to within 2.0% for field sizes ranging from 2 × 2 to 40 × 40 cm2. Measured and MC profiles show agreement in the 50% field width and the 80%‐20% penumbra region to within 1.3 mm for all square field sizes. MC OFs for the 2 to 40 cm2 square fields agree with measurement to within 1.6%. Verification of VMAT SABR lung, liver, and vertebra plans demonstrate that measured and MC ion chamber doses agree within 0.6% for the 6X beam and within 2.0% for the 6X FFF beam. A 3D gamma factor analysis demonstrates that for the 6X beam, > 99% of voxels meet the pass criteria (3%/3 mm). For the 6X FFF beam, > 94% of voxels meet this criteria. The TrueBeam accelerator delivering 6X and 6X FFF beams with the HD120 MLC can be modeled in Monte Carlo to provide an independent 3D dose calculation for clinical VMAT plans. This quality assurance tool has been used clinically to verify over 140 6X and 16 6X FFF TrueBeam treatment plans. PACS number: 87.55.K‐ PMID:24892341
TU-FG-201-09: Predicting Accelerator Dysfunction
DOE Office of Scientific and Technical Information (OSTI.GOV)
Able, C; Nguyen, C; Baydush, A
Purpose: To develop an integrated statistical process control (SPC) framework using digital performance and component data accumulated within the accelerator system that can detect dysfunction prior to unscheduled downtime. Methods: Seven digital accelerators were monitored for twelve to 18 months. The accelerators were operated in a ‘run to failure mode’ with the individual institutions determining when service would be initiated. Institutions were required to submit detailed service reports. Trajectory and text log files resulting from a robust daily VMAT QA delivery were decoded and evaluated using Individual and Moving Range (I/MR) control charts. The SPC evaluation was presented in amore » customized dashboard interface that allows the user to review 525 monitored parameters (480 MLC parameters). Chart limits were calculated using a hybrid technique that includes the standard SPC 3σ limits and an empirical factor based on the parameter/system specification. The individual (I) grand mean values and control limit ranges of the I/MR charts of all accelerators were compared using statistical (ranked analysis of variance (ANOVA)) and graphical analyses to determine consistency of operating parameters. Results: When an alarm or warning was directly connected to field service, process control charts predicted dysfunction consistently on beam generation related parameters (BGP)– RF Driver Voltage, Gun Grid Voltage, and Forward Power (W); beam uniformity parameters – angle and position steering coil currents; and Gantry position accuracy parameter: cross correlation max-value. Control charts for individual MLC – cross correlation max-value/position detected 50% to 60% of MLCs serviced prior to dysfunction or failure. In general, non-random changes were detected 5 to 80 days prior to a service intervention. The ANOVA comparison of BGP determined that each accelerator parameter operated at a distinct value. Conclusion: The SPC framework shows promise. Long term monitoring coordinated with service will be required to definitively determine the effectiveness of the model. Varian Medical System, Inc. provided funding in support of the research presented.« less
SU-E-T-142: Automatic Linac Log File: Analysis and Reporting
DOE Office of Scientific and Technical Information (OSTI.GOV)
Gainey, M; Rothe, T
Purpose: End to end QA for IMRT/VMAT is time consuming. Automated linac log file analysis and recalculation of daily recorded fluence, and hence dose, distribution bring this closer. Methods: Matlab (R2014b, Mathworks) software was written to read in and analyse IMRT/VMAT trajectory log files (TrueBeam 1.5, Varian Medical Systems) overnight, and are archived on a backed-up network drive (figure). A summary report (PDF) is sent by email to the duty linac physicist. A structured summary report (PDF) for each patient is automatically updated for embedding into the R&V system (Mosaiq 2.5, Elekta AG). The report contains cross-referenced hyperlinks to easemore » navigation between treatment fractions. Gamma analysis can be performed on planned (DICOM RTPlan) and treated (trajectory log) fluence distributions. Trajectory log files can be converted into RTPlan files for dose distribution calculation (Eclipse, AAA10.0.28, VMS). Results: All leaf positions are within +/−0.10mm: 57% within +/−0.01mm; 89% within 0.05mm. Mean leaf position deviation is 0.02mm. Gantry angle variations lie in the range −0.1 to 0.3 degrees, mean 0.04 degrees. Fluence verification shows excellent agreement between planned and treated fluence. Agreement between planned and treated dose distribution, the derived from log files, is very good. Conclusion: Automated log file analysis is a valuable tool for the busy physicist, enabling potential treated fluence distribution errors to be quickly identified. In the near future we will correlate trajectory log analysis with routine IMRT/VMAT QA analysis. This has the potential to reduce, but not eliminate, the QA workload.« less
46 CFR Appendix A to Part 530 - Instructions for the Filing of Service Contracts
Code of Federal Regulations, 2011 CFR
2011-10-01
... file service contracts. BTCL will direct OIRM to provide approved filers with a log-on ID and password. Filers who wish a third party (publisher) to file their service contracts must so indicate on Form FMC-83... home page, http://www.fmc.gov. A. Registration, Log-on ID and Password To register for filing, a...
46 CFR Appendix A to Part 530 - Instructions for the Filing of Service Contracts
Code of Federal Regulations, 2010 CFR
2010-10-01
... file service contracts. BTCL will direct OIRM to provide approved filers with a log-on ID and password. Filers who wish a third party (publisher) to file their service contracts must so indicate on Form FMC-83... home page, http://www.fmc.gov. A. Registration, Log-on ID and Password To register for filing, a...
DOE Office of Scientific and Technical Information (OSTI.GOV)
Tang, S; Ho, M; Chen, C
Purpose: The use of log files to perform patient specific quality assurance for both protons and IMRT has been established. Here, we extend that approach to a proprietary log file format and compare our results to measurements in phantom. Our goal was to generate a system that would permit gross errors to be found within 3 fractions until direct measurements. This approach could eventually replace direct measurements. Methods: Spot scanning protons pass through multi-wire ionization chambers which provide information about the charge, location, and size of each delivered spot. We have generated a program that calculates the dose in phantommore » from these log files and compares the measurements with the plan. The program has 3 different spot shape models: single Gaussian, double Gaussian and the ASTROID model. The program was benchmarked across different treatment sites for 23 patients and 74 fields. Results: The dose calculated from the log files were compared to those generate by the treatment planning system (Raystation). While the dual Gaussian model often gave better agreement, overall, the ASTROID model gave the most consistent results. Using a 5%–3 mm gamma with a 90% passing criteria and excluding doses below 20% of prescription all patient samples passed. However, the degree of agreement of the log file approach was slightly worse than that of the chamber array measurement approach. Operationally, this implies that if the beam passes the log file model, it should pass direct measurement. Conclusion: We have established and benchmarked a model for log file QA in an IBA proteus plus system. The choice of optimal spot model for a given class of patients may be affected by factors such as site, field size, and range shifter and will be investigated further.« less
An EXCEL macro for importing log ASCII standard (LAS) files into EXCEL worksheets
NASA Astrophysics Data System (ADS)
Özkaya, Sait Ismail
1996-02-01
An EXCEL 5.0 macro is presented for converting a LAS text file into an EXCEL worksheet. Although EXCEL has commands for importing text files and parsing text lines, LAS files must be decoded line-by-line because three different delimiters are used to separate fields of differing length. The macro is intended to eliminate manual decoding of LAS version 2.0. LAS is a floppy disk format for storage and transfer of log data as text files. LAS was proposed by the Canadian Well Logging Society. The present EXCEL macro decodes different sections of a LAS file, separates, and places the fields into different columns of an EXCEL worksheet. To import a LAS file into EXCEL without errors, the file must not contain any unrecognized symbols, and the data section must be the last section. The program does not check for the presence of mandatory sections or fields as required by LAS rules. Once a file is incorporated into EXCEL, mandatory sections and fields may be inspected visually.
SU-E-T-184: Clinical VMAT QA Practice Using LINAC Delivery Log Files
DOE Office of Scientific and Technical Information (OSTI.GOV)
Johnston, H; Jacobson, T; Gu, X
2015-06-15
Purpose: To evaluate the accuracy of volumetric modulated arc therapy (VMAT) treatment delivery dose clouds by comparing linac log data to doses measured using an ionization chamber and film. Methods: A commercial IMRT quality assurance (QA) process utilizing a DICOM-RT framework was tested for clinical practice using 30 prostate and 30 head and neck VMAT plans. Delivered 3D VMAT dose distributions were independently checked using a PinPoint ionization chamber and radiographic film in a solid water phantom. DICOM RT coordinates were used to extract the corresponding point and planar doses from 3D log file dose distributions. Point doses were evaluatedmore » by computing the percent error between log file and chamber measured values. A planar dose evaluation was performed for each plan using a 2D gamma analysis with 3% global dose difference and 3 mm isodose point distance criteria. The same analysis was performed to compare treatment planning system (TPS) doses to measured values to establish a baseline assessment of agreement. Results: The mean percent error between log file and ionization chamber dose was 1.0%±2.1% for prostate VMAT plans and −0.2%±1.4% for head and neck plans. The corresponding TPS calculated and measured ionization chamber values agree within 1.7%±1.6%. The average 2D gamma passing rates for the log file comparison to film are 98.8%±1.0% and 96.2%±4.2% for the prostate and head and neck plans, respectively. The corresponding passing rates for the TPS comparison to film are 99.4%±0.5% and 93.9%±5.1%. Overall, the point dose and film data indicate that log file determined doses are in excellent agreement with measured values. Conclusion: Clinical VMAT QA practice using LINAC treatment log files is a fast and reliable method for patient-specific plan evaluation.« less
NASA Astrophysics Data System (ADS)
Xu, Jun; Papanikolaou, Nikos; Shi, Chengyu; Jiang, Steve B.
2009-08-01
Synchronized moving aperture radiation therapy (SMART) has been proposed to account for tumor motions during radiotherapy in prior work. The basic idea of SMART is to synchronize the moving radiation beam aperture formed by a dynamic multileaf collimator (DMLC) with the tumor motion induced by respiration. In this paper, a two-dimensional (2D) superimposing leaf sequencing method is presented for SMART. A leaf sequence optimization strategy was generated to assure the SMART delivery under realistic delivery conditions. The study of delivery performance using the Varian LINAC and the Millennium DMLC showed that clinical factors such as collimator angle, dose rate, initial phase and machine tolerance affect the delivery accuracy and efficiency. An in-house leaf sequencing software was developed to implement the 2D superimposing leaf sequencing method and optimize the motion-corrected leaf sequence under realistic clinical conditions. The analysis of dynamic log (Dynalog) files showed that optimization of the leaf sequence for various clinical factors can avoid beam hold-offs which break the synchronization of SMART and fail the SMART dose delivery. Through comparison between the simulated delivered fluence map and the planed fluence map, it was shown that the motion-corrected leaf sequence can greatly reduce the dose error.
Xu, Jun; Papanikolaou, Nikos; Shi, Chengyu; Jiang, Steve B
2009-08-21
Synchronized moving aperture radiation therapy (SMART) has been proposed to account for tumor motions during radiotherapy in prior work. The basic idea of SMART is to synchronize the moving radiation beam aperture formed by a dynamic multileaf collimator (DMLC) with the tumor motion induced by respiration. In this paper, a two-dimensional (2D) superimposing leaf sequencing method is presented for SMART. A leaf sequence optimization strategy was generated to assure the SMART delivery under realistic delivery conditions. The study of delivery performance using the Varian LINAC and the Millennium DMLC showed that clinical factors such as collimator angle, dose rate, initial phase and machine tolerance affect the delivery accuracy and efficiency. An in-house leaf sequencing software was developed to implement the 2D superimposing leaf sequencing method and optimize the motion-corrected leaf sequence under realistic clinical conditions. The analysis of dynamic log (Dynalog) files showed that optimization of the leaf sequence for various clinical factors can avoid beam hold-offs which break the synchronization of SMART and fail the SMART dose delivery. Through comparison between the simulated delivered fluence map and the planed fluence map, it was shown that the motion-corrected leaf sequence can greatly reduce the dose error.
Andruchov, Oleg; Andruchova, Olena; Wang, Yishu; Galler, Stefan
2006-02-15
Cross-bridge kinetics underlying stretch-induced force transients was studied in fibres with different myosin light chain (MLC) isoforms from skeletal muscles of rabbit and rat. The force transients were induced by stepwise stretches (< 0.3% of fibre length) applied on maximally Ca2+-activated skinned fibres. Fast fibre types IIB, IID (or IIX) and IIA and the slow fibre type I containing the myosin heavy chain isoforms MHC-IIb, MHC-IId (or MHC-IIx), MHC-IIa and MHC-I, respectively, were investigated. The MLC isoform content varied within fibre types. Fast fibre types contained the fast regulatory MLC isoform MLC2f and different proportions of the fast alkali MLC isoforms MLC1f and MLC3f. Type I fibres contained the slow regulatory MLC isoform MLC2s and the slow alkali MLC isoform MLC1s. Slow MLC isoforms were also present in several type IIA fibres. The kinetics of force transients differed by a factor of about 30 between fibre types (order from fastest to slowest kinetics: IIB > IID > IIA > I). The kinetics of the force transients was not dependent on the relative content of MLC1f and MLC3f. Type IIA fibres containing fast and slow MLC isoforms were about 1.2 times slower than type IIA fibres containing only fast MLC isoforms. We conclude that while the cross-bridge kinetics is mainly determined by the MHC isoforms present, it is affected by fast and slow MLC isoforms but not by the relative content of MLC1f and MLC3f. Thus, the physiological role of fast and slow MLC isoforms in type IIA fibres is a fine-tuning of the cross-bridge kinetics.
Megalencephalic leukoencephalopathy with subcortical cysts: Characterization of disease variants.
Hamilton, Eline M C; Tekturk, Pinar; Cialdella, Fia; van Rappard, Diane F; Wolf, Nicole I; Yalcinkaya, Cengiz; Çetinçelik, Ümran; Rajaee, Ahmad; Kariminejad, Ariana; Paprocka, Justyna; Yapici, Zuhal; Bošnjak, Vlatka Mejaški; van der Knaap, Marjo S
2018-04-17
To provide an overview of clinical and MRI characteristics of the different variants of the leukodystrophy megalencephalic leukoencephalopathy with subcortical cysts (MLC) and identify possible differentiating features. We performed an international multi-institutional, cross-sectional observational study of the clinical and MRI characteristics in patients with genetically confirmed MLC. Clinical information was obtained by questionnaires for physicians and retrospective chart review. We included 204 patients with classic MLC, 187 of whom had recessive mutations in MLC1 (MLC1 variant) and 17 in GLIALCAM (MLC2A variant) and 38 patients with remitting MLC caused by dominant GLIALCAM mutations (MLC2B variant). We observed a relatively wide variability in neurologic disability among patients with classic MLC. No clinical differences could be identified between patients with MLC1 and MLC2A. Patients with MLC2B invariably had a milder phenotype with preservation of motor function, while intellectual disability and autism were relatively frequent. Systematic MRI review revealed no MRI features that distinguish between MLC1 and MLC2A. Radiologic improvement was observed in all patients with MLC2B and also in 2 patients with MLC1. In MRIs obtained in the early disease stage, absence of signal abnormalities of the posterior limb of the internal capsule and cerebellar white matter and presence of only rarefied subcortical white matter instead of true subcortical cysts were suggestive of MLC2B. Clinical and MRI features did not distinguish between classic MLC with MLC1 or GLIALCAM mutations. Absence of signal abnormalities of the internal capsule and cerebellar white matter are MRI findings that point to the remitting phenotype. Copyright © 2018 The Author(s). Published by Wolters Kluwer Health, Inc. on behalf of the American Academy of Neurology.
Park, So-Yeon; Kim, Il Han; Ye, Sung-Joon; Carlson, Joel; Park, Jong Min
2014-11-01
Texture analysis on fluence maps was performed to evaluate the degree of modulation for volumetric modulated arc therapy (VMAT) plans. A total of six textural features including angular second moment, inverse difference moment, contrast, variance, correlation, and entropy were calculated for fluence maps generated from 20 prostate and 20 head and neck VMAT plans. For each of the textural features, particular displacement distances (d) of 1, 5, and 10 were adopted. To investigate the deliverability of each VMAT plan, gamma passing rates of pretreatment quality assurance, and differences in modulating parameters such as multileaf collimator (MLC) positions, gantry angles, and monitor units at each control point between VMAT plans and dynamic log files registered by the Linac control system during delivery were acquired. Furthermore, differences between the original VMAT plan and the plan reconstructed from the dynamic log files were also investigated. To test the performance of the textural features as indicators for the modulation degree of VMAT plans, Spearman's rank correlation coefficients (rs) with the plan deliverability were calculated. For comparison purposes, conventional modulation indices for VMAT including the modulation complexity score for VMAT, leaf travel modulation complexity score, and modulation index supporting station parameter optimized radiation therapy (MISPORT) were calculated, and their correlations were analyzed in the same way. There was no particular textural feature which always showed superior correlations with every type of plan deliverability. Considering the results comprehensively, contrast (d = 1) and variance (d = 1) generally showed considerable correlations with every type of plan deliverability. These textural features always showed higher correlations to the plan deliverability than did the conventional modulation indices, except in the case of modulating parameter differences. The rs values of contrast to the global gamma passing rates with criteria of 2%/2 mm, 2%/1 mm, and 1%/2 mm were 0.536, 0.473, and 0.718, respectively. The respective values for variance were 0.551, 0.481, and 0.688. In the case of local gamma passing rates, the rs values of contrast were 0.547, 0.578, and 0.620, respectively, and those of variance were 0.519, 0.527, and 0.569. All of the rs values in those cases were statistically significant (p < 0.003). In the cases of global and local gamma passing rates, MISPORT showed the highest correlations among the conventional modulation indices. For global passing rates, rs values of MISPORT were -0.420, -0.330, and -0.632, respectively, and those for local passing rates were -0.455, -0.490 and -0.502. The values of rs of contrast, variance, and MISPORT with the MLC errors were -0.863, -0.828, and 0.795, respectively, all with statistical significances (p < 0.001). The correlations with statistical significances between variance and dose-volumetric differences were observed more frequently than the others. The contrast (d = 1) and variance (d = 1) calculated from fluence maps of VMAT plans showed considerable correlations with the plan deliverability, indicating their potential use as indicators for assessing the degree of modulation of VMAT plans. Both contrast and variance consistently showed better performance than the conventional modulation indices for VMAT.
Jorde, Ulrich P; Aaronson, Keith D; Najjar, Samer S; Pagani, Francis D; Hayward, Christopher; Zimpfer, Daniel; Schlöglhofer, Thomas; Pham, Duc T; Goldstein, Daniel J; Leadley, Katrin; Chow, Ming-Jay; Brown, Michael C; Uriel, Nir
2015-11-01
The study sought to characterize patterns in the HeartWare (HeartWare Inc., Framingham, Massachusetts) ventricular assist device (HVAD) log files associated with successful medical treatment of device thrombosis. Device thrombosis is a serious adverse event for mechanical circulatory support devices and is often preceded by increased power consumption. Log files of the pump power are easily accessible on the bedside monitor of HVAD patients and may allow early diagnosis of device thrombosis. Furthermore, analysis of the log files may be able to predict the success rate of thrombolysis or the need for pump exchange. The log files of 15 ADVANCE trial patients (algorithm derivation cohort) with 16 pump thrombus events treated with tissue plasminogen activator (tPA) were assessed for changes in the absolute and rate of increase in power consumption. Successful thrombolysis was defined as a clinical resolution of pump thrombus including normalization of power consumption and improvement in biochemical markers of hemolysis. Significant differences in log file patterns between successful and unsuccessful thrombolysis treatments were verified in 43 patients with 53 pump thrombus events implanted outside of clinical trials (validation cohort). The overall success rate of tPA therapy was 57%. Successful treatments had significantly lower measures of percent of expected power (130.9% vs. 196.1%, p = 0.016) and rate of increase in power (0.61 vs. 2.87, p < 0.0001). Medical therapy was successful in 77.7% of the algorithm development cohort and 81.3% of the validation cohort when the rate of power increase and percent of expected power values were <1.25% and 200%, respectively. Log file parameters can potentially predict the likelihood of successful tPA treatments and if validated prospectively, could substantially alter the approach to thrombus management. Copyright © 2015 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Huang, Min; Liu, Zhaoqing; Qiao, Liyan
2014-10-10
While the NAND flash memory is widely used as the storage medium in modern sensor systems, the aggressive shrinking of process geometry and an increase in the number of bits stored in each memory cell will inevitably degrade the reliability of NAND flash memory. In particular, it's critical to enhance metadata reliability, which occupies only a small portion of the storage space, but maintains the critical information of the file system and the address translations of the storage system. Metadata damage will cause the system to crash or a large amount of data to be lost. This paper presents Asymmetric Programming, a highly reliable metadata allocation strategy for MLC NAND flash memory storage systems. Our technique exploits for the first time the property of the multi-page architecture of MLC NAND flash memory to improve the reliability of metadata. The basic idea is to keep metadata in most significant bit (MSB) pages which are more reliable than least significant bit (LSB) pages. Thus, we can achieve relatively low bit error rates for metadata. Based on this idea, we propose two strategies to optimize address mapping and garbage collection. We have implemented Asymmetric Programming on a real hardware platform. The experimental results show that Asymmetric Programming can achieve a reduction in the number of page errors of up to 99.05% with the baseline error correction scheme.
Huang, Min; Liu, Zhaoqing; Qiao, Liyan
2014-01-01
While the NAND flash memory is widely used as the storage medium in modern sensor systems, the aggressive shrinking of process geometry and an increase in the number of bits stored in each memory cell will inevitably degrade the reliability of NAND flash memory. In particular, it's critical to enhance metadata reliability, which occupies only a small portion of the storage space, but maintains the critical information of the file system and the address translations of the storage system. Metadata damage will cause the system to crash or a large amount of data to be lost. This paper presents Asymmetric Programming, a highly reliable metadata allocation strategy for MLC NAND flash memory storage systems. Our technique exploits for the first time the property of the multi-page architecture of MLC NAND flash memory to improve the reliability of metadata. The basic idea is to keep metadata in most significant bit (MSB) pages which are more reliable than least significant bit (LSB) pages. Thus, we can achieve relatively low bit error rates for metadata. Based on this idea, we propose two strategies to optimize address mapping and garbage collection. We have implemented Asymmetric Programming on a real hardware platform. The experimental results show that Asymmetric Programming can achieve a reduction in the number of page errors of up to 99.05% with the baseline error correction scheme. PMID:25310473
An experimental comparison of conventional two-bank and novel four-bank dynamic MLC tracking.
Davies, G A; Clowes, P; McQuaid, D; Evans, P M; Webb, S; Poludniowski, G
2013-03-07
The AccuLeaf mMLC featuring four multileaf-collimator (MLC) banks has been used for the first time for an experimental comparison of conventional two-bank with novel four-bank dynamic MLC tracking of a two-dimensional sinusoidal respiratory motion. This comparison was performed for a square aperture, and for three conformal treatment apertures from clinical radiotherapy lung cancer patients. The system latency of this prototype tracking system was evaluated and found to be 1.0 s and the frequency at which MLC positions could be updated, 1 Hz, and therefore accurate MLC tracking of irregular patient motion would be difficult with the system in its current form. The MLC leaf velocity required for two-bank-MLC and four-bank-MLC tracking was evaluated for the apertures studied and a substantial decrease was found in the maximum MLC velocity required when four-banks were used for tracking rather than two. A dosimetric comparison of the two techniques was also performed and minimal difference was found between two-bank-MLC and four-bank-MLC tracking. The use of four MLC banks for dynamic MLC tracking is shown to be potentially advantageous for increasing the delivery efficiency compared with two-bank-MLC tracking where difficulties are encountered if large leaf shifts are required to track motion perpendicular to the direction of leaf travel.
Létourneau, Daniel; Wang, An; Amin, Md Nurul; Pearce, Jim; McNiven, Andrea; Keller, Harald; Norrlinger, Bernhard; Jaffray, David A
2014-12-01
High-quality radiation therapy using highly conformal dose distributions and image-guided techniques requires optimum machine delivery performance. In this work, a monitoring system for multileaf collimator (MLC) performance, integrating semiautomated MLC quality control (QC) tests and statistical process control tools, was developed. The MLC performance monitoring system was used for almost a year on two commercially available MLC models. Control charts were used to establish MLC performance and assess test frequency required to achieve a given level of performance. MLC-related interlocks and servicing events were recorded during the monitoring period and were investigated as indicators of MLC performance variations. The QC test developed as part of the MLC performance monitoring system uses 2D megavoltage images (acquired using an electronic portal imaging device) of 23 fields to determine the location of the leaves with respect to the radiation isocenter. The precision of the MLC performance monitoring QC test and the MLC itself was assessed by detecting the MLC leaf positions on 127 megavoltage images of a static field. After initial calibration, the MLC performance monitoring QC test was performed 3-4 times/week over a period of 10-11 months to monitor positional accuracy of individual leaves for two different MLC models. Analysis of test results was performed using individuals control charts per leaf with control limits computed based on the measurements as well as two sets of specifications of ± 0.5 and ± 1 mm. Out-of-specification and out-of-control leaves were automatically flagged by the monitoring system and reviewed monthly by physicists. MLC-related interlocks reported by the linear accelerator and servicing events were recorded to help identify potential causes of nonrandom MLC leaf positioning variations. The precision of the MLC performance monitoring QC test and the MLC itself was within ± 0.22 mm for most MLC leaves and the majority of the apparent leaf motion was attributed to beam spot displacements between irradiations. The MLC QC test was performed 193 and 162 times over the monitoring period for the studied units and recalibration had to be repeated up to three times on one of these units. For both units, rate of MLC interlocks was moderately associated with MLC servicing events. The strongest association with the MLC performance was observed between the MLC servicing events and the total number of out-of-control leaves. The average elapsed time for which the number of out-of-specification or out-of-control leaves was within a given performance threshold was computed and used to assess adequacy of MLC test frequency. A MLC performance monitoring system has been developed and implemented to acquire high-quality QC data at high frequency. This is enabled by the relatively short acquisition time for the images and automatic image analysis. The monitoring system was also used to record and track the rate of MLC-related interlocks and servicing events. MLC performances for two commercially available MLC models have been assessed and the results support monthly test frequency for widely accepted ± 1 mm specifications. Higher QC test frequency is however required to maintain tighter specification and in-control behavior.
Radio frequency noise from an MLC: a feasibility study of the use of an MLC for linac-MR systems.
Lamey, M; Yun, J; Burke, B; Rathee, S; Fallone, B G
2010-02-21
Currently several groups are actively researching the integration of a megavoltage teletherapy unit with magnetic resonance (MR) imaging for real-time image-guided radiotherapy. The use of a multileaf collimator (MLC) for intensity-modulated radiotherapy for linac-MR units must be investigated. The MLC itself will likely reside in the fringe field of the MR and the motors will produce radio frequency (RF) noise. The RF noise power spectral density from a Varian 52-leaf MLC motor, a Varian Millennium MLC motor and a brushless fan motor has been measured as a function of the applied magnetic field using a near field probe set. For the Varian 52-leaf MLC system, the RF noise produced by 13 of 52 motors is studied as a function of distance from the MLC. Data are reported in the frequency range suitable for 0.2-1.5 T linac-MR systems. Below 40 MHz the Millennium MLC motor tested showed more noise than the Varian 52-leaf motor or the brushless fan motor. The brushless motor showed a small dependence on the applied magnetic field. Images of a phantom were taken by the prototype linac-MR system with the MLC placed in close proximity to the magnet. Several orientations of the MLC in both shielded and non-shielded configurations were studied. For the case of a non-shielded MLC and associated cables, the signal-to-noise ratio (SNR) was reduced when 13 of 52 MLC leaves were moved during imaging. When the MLC and associated cables were shielded, the measured SNR of the images with 13 MLC leaves moving was experimentally the same as the SNR of the stationary MLC image. When the MLC and cables are shielded, subtraction images acquired with and without MLC motion contains no systematic signal. This study illustrates that the small RF noise produced by functioning MLC motors can be effectively shielded to avoid SNR degradation. A functioning MLC can be incorporated into a linac-MR unit.
Radio frequency noise from an MLC: a feasibility study of the use of an MLC for linac-MR systems
Lamey, M; Yun, J; Burke, B; Rathee, S; Fallone, B G
2010-01-01
Currently several groups are actively researching the integration of a megavoltage teletherapy unit with magnetic resonance (MR) imaging for real-time image-guided radiotherapy. The use of a multileaf collimator (MLC) for intensity-modulated radiotherapy for linac-MR units must be investigated. The MLC itself will likely reside in the fringe field of the MR and the motors will produce radio frequency (RF) noise. The RF noise power spectral density from a Varian 52-leaf MLC motor, a Varian Millennium MLC motor and a brushless fan motor has been measured as a function of the applied magnetic field using a near field probe set. For the Varian 52-leaf MLC system, the RF noise produced by 13 of 52 motors is studied as a function of distance from the MLC. Data are reported in the frequency range suitable for 0.2–1.5 T linac-MR systems. Below 40 MHz the Millennium MLC motor tested showed more noise than the Varian 52-leaf motor or the brushless fan motor. The brushless motor showed a small dependence on the applied magnetic field. Images of a phantom were taken by the prototype linac-MR system with the MLC placed in close proximity to the magnet. Several orientations of the MLC in both shielded and non-shielded configurations were studied. For the case of a non-shielded MLC and associated cables, the signal-to-noise ratio (SNR) was reduced when 13 of 52 MLC leaves were moved during imaging. When the MLC and associated cables were shielded, the measured SNR of the images with 13 MLC leaves moving was experimentally the same as the SNR of the stationary MLC image. When the MLC and cables are shielded, subtraction images acquired with and without MLC motion contains no systematic signal. This study illustrates that the small RF noise produced by functioning MLC motors can be effectively shielded to avoid SNR degradation. A functioning MLC can be incorporated into a linac-MR unit. PMID:20090187
Log-less metadata management on metadata server for parallel file systems.
Liao, Jianwei; Xiao, Guoqiang; Peng, Xiaoning
2014-01-01
This paper presents a novel metadata management mechanism on the metadata server (MDS) for parallel and distributed file systems. In this technique, the client file system backs up the sent metadata requests, which have been handled by the metadata server, so that the MDS does not need to log metadata changes to nonvolatile storage for achieving highly available metadata service, as well as better performance improvement in metadata processing. As the client file system backs up certain sent metadata requests in its memory, the overhead for handling these backup requests is much smaller than that brought by the metadata server, while it adopts logging or journaling to yield highly available metadata service. The experimental results show that this newly proposed mechanism can significantly improve the speed of metadata processing and render a better I/O data throughput, in contrast to conventional metadata management schemes, that is, logging or journaling on MDS. Besides, a complete metadata recovery can be achieved by replaying the backup logs cached by all involved clients, when the metadata server has crashed or gone into nonoperational state exceptionally.
Log-Less Metadata Management on Metadata Server for Parallel File Systems
Xiao, Guoqiang; Peng, Xiaoning
2014-01-01
This paper presents a novel metadata management mechanism on the metadata server (MDS) for parallel and distributed file systems. In this technique, the client file system backs up the sent metadata requests, which have been handled by the metadata server, so that the MDS does not need to log metadata changes to nonvolatile storage for achieving highly available metadata service, as well as better performance improvement in metadata processing. As the client file system backs up certain sent metadata requests in its memory, the overhead for handling these backup requests is much smaller than that brought by the metadata server, while it adopts logging or journaling to yield highly available metadata service. The experimental results show that this newly proposed mechanism can significantly improve the speed of metadata processing and render a better I/O data throughput, in contrast to conventional metadata management schemes, that is, logging or journaling on MDS. Besides, a complete metadata recovery can be achieved by replaying the backup logs cached by all involved clients, when the metadata server has crashed or gone into nonoperational state exceptionally. PMID:24892093
Williams, Lester J.; Raines, Jessica E.; Lanning, Amanda E.
2013-04-04
A database of borehole geophysical logs and other types of data files were compiled as part of ongoing studies of water availability and assessment of brackish- and saline-water resources. The database contains 4,883 logs from 1,248 wells in Florida, Georgia, Alabama, South Carolina, and from a limited number of offshore wells of the eastern Gulf of Mexico and the Atlantic Ocean. The logs can be accessed through a download directory organized by state and county for onshore wells and in a single directory for the offshore wells. A flat file database is provided that lists the wells, their coordinates, and the file listings.
NASA Astrophysics Data System (ADS)
Madiraju, Praveen; Zhang, Yanqing
2002-03-01
When a user logs in to a website, behind the scenes the user leaves his/her impressions, usage patterns and also access patterns in the web servers log file. A web usage mining agent can analyze these web logs to help web developers to improve the organization and presentation of their websites. They can help system administrators in improving the system performance. Web logs provide invaluable help in creating adaptive web sites and also in analyzing the network traffic analysis. This paper presents the design and implementation of a Web usage mining agent for digging in to the web log files.
Ryder, Robert T.; Trippi, Michael H.; Ruppert, Leslie F.; Ryder, Robert T.
2014-01-01
The appendixes in chapters E.4.1 and E.4.2 include (1) Log ASCII Standard (LAS) files, which encode gamma-ray, neutron, density, and other logs in text files that can be used by most well-logging software programs; and (2) graphic well-log traces. In the appendix to chapter E.4.1, the well-log traces are accompanied by lithologic descriptions with formation tops.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Létourneau, Daniel, E-mail: daniel.letourneau@rmp.uh.on.ca; McNiven, Andrea; Keller, Harald
2014-12-15
Purpose: High-quality radiation therapy using highly conformal dose distributions and image-guided techniques requires optimum machine delivery performance. In this work, a monitoring system for multileaf collimator (MLC) performance, integrating semiautomated MLC quality control (QC) tests and statistical process control tools, was developed. The MLC performance monitoring system was used for almost a year on two commercially available MLC models. Control charts were used to establish MLC performance and assess test frequency required to achieve a given level of performance. MLC-related interlocks and servicing events were recorded during the monitoring period and were investigated as indicators of MLC performance variations. Methods:more » The QC test developed as part of the MLC performance monitoring system uses 2D megavoltage images (acquired using an electronic portal imaging device) of 23 fields to determine the location of the leaves with respect to the radiation isocenter. The precision of the MLC performance monitoring QC test and the MLC itself was assessed by detecting the MLC leaf positions on 127 megavoltage images of a static field. After initial calibration, the MLC performance monitoring QC test was performed 3–4 times/week over a period of 10–11 months to monitor positional accuracy of individual leaves for two different MLC models. Analysis of test results was performed using individuals control charts per leaf with control limits computed based on the measurements as well as two sets of specifications of ±0.5 and ±1 mm. Out-of-specification and out-of-control leaves were automatically flagged by the monitoring system and reviewed monthly by physicists. MLC-related interlocks reported by the linear accelerator and servicing events were recorded to help identify potential causes of nonrandom MLC leaf positioning variations. Results: The precision of the MLC performance monitoring QC test and the MLC itself was within ±0.22 mm for most MLC leaves and the majority of the apparent leaf motion was attributed to beam spot displacements between irradiations. The MLC QC test was performed 193 and 162 times over the monitoring period for the studied units and recalibration had to be repeated up to three times on one of these units. For both units, rate of MLC interlocks was moderately associated with MLC servicing events. The strongest association with the MLC performance was observed between the MLC servicing events and the total number of out-of-control leaves. The average elapsed time for which the number of out-of-specification or out-of-control leaves was within a given performance threshold was computed and used to assess adequacy of MLC test frequency. Conclusions: A MLC performance monitoring system has been developed and implemented to acquire high-quality QC data at high frequency. This is enabled by the relatively short acquisition time for the images and automatic image analysis. The monitoring system was also used to record and track the rate of MLC-related interlocks and servicing events. MLC performances for two commercially available MLC models have been assessed and the results support monthly test frequency for widely accepted ±1 mm specifications. Higher QC test frequency is however required to maintain tighter specification and in-control behavior.« less
Forskolin induces myosin light chain dephosphorylation in bovine trabecular meshwork cells.
Ramachandran, Charanya; Satpathy, Minati; Mehta, Dolly; Srinivas, Sangly P
2008-02-01
Enhanced contractility of the actin cytoskeleton in trabecular meshwork (TM) cells is implicated in increased resistance to aqueous humor outflow. In this study, we have investigated effects of forskolin, which is known to elevate cAMP and also enhance aqueous humor outflow, on myosin light chain (MLC) phosphorylation, a biochemical marker of actin contractility. Experiments were performed using cultured bovine TM cells. Phosphorylated MLC (pMLC), expressed as the % of untreated cells, was assessed by urea-glycerol gel electrophoresis and Western blotting. RhoA activity was determined by affinity precipitation of RhoA-GTP to RhoA binding domain of an effector of RhoA. Intracellular cAMP levels were measured by ELISA. Exposure to LPA (lysophosphatidic acid) led to increased MLC phosphorylation (LPA: pMLC=133%) and activation of RhoA. These responses of LPA were suppressed by co-treatment with forskolin (LPA+forskolin: pMLC=88%). Similarly, ET-1 and nocodazole-induced MLC phosphorylation (ET-1: pMLC=145%; nocodazole: pMLC=145%) as well as RhoA activation were suppressed by co-treatment with forskolin (ET-1+forskolin: pMLC=99%; nocodazole+forskolin: pMLC=107%). Exposure to forskolin alone led to MLC dephosphorylation (pMLC=68%). Forskolin alone led to a 4-fold increase in cAMP levels. This increase was not affected when co-treated with LPA or ET-1. Forskolin prevents MLC phosphorylation induced by LPA, ET-1, and nocodazole through inhibition of RhoA-Rho kinase axis. MLC dephosphorylation and consequent relaxation of actin cytoskeleton in TM cells presumably underlies the increased outflow facility reported in response to forskolin.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Xu, Z; Wang, I; Yao, R
Purpose: This study is to use plan parameters optimization (Dose rate, collimator angle, couch angle, initial starting phase) to improve the performance of conformal arc radiotherapy plans with motion tracking by increasing the plan performance score (PPS). Methods: Two types of 3D conformal arc plans were created based on QUASAR respiratory motion phantom with spherical and cylindrical targets. Sinusoidal model was applied to the MLC leaves to generate motion tracking plans. A MATLAB program was developed to calculate PPS of each plan (ranges from 0–1) and optimize plan parameters. We first selected the dose rate for motion tracking plans andmore » then used simulated annealing algorithm to search for the combination of the other parameters that resulted in the plan of the maximal PPS. The optimized motion tracking plan was delivered by Varian Truebeam Linac. In-room cameras and stopwatch were used for starting phase selection and synchronization between phantom motion and plan delivery. Gaf-EBT2 dosimetry films were used to measure the dose delivered to the target in QUASAR phantom. Dose profiles and Truebeam trajectory log files were used for plan delivery performance evaluation. Results: For spherical target, the maximal PPS (PPSsph) of the optimized plan was 0.79: (Dose rate: 500MU/min, Collimator: 90°, Couch: +10°, starting phase: 0.83π). For cylindrical target, the maximal PPScyl was 0.75 (Dose rate: 300MU/min, Collimator: 87°, starting phase: 0.97π) with couch at 0°. Differences of dose profiles between motion tracking plans (with the maximal and the minimal PPS) and 3D conformal plans were as follows: PPSsph=0.79: %ΔFWHM: 8.9%, %Dmax: 3.1%; PPSsph=0.52: %ΔFWHM: 10.4%, %Dmax: 6.1%. PPScyl=0.75: %ΔFWHM: 4.7%, %Dmax: 3.6%; PPScyl=0.42: %ΔFWHM: 12.5%, %Dmax: 9.6%. Conclusion: By achieving high plan performance score through parameters optimization, we can improve target dose conformity of motion tracking plan by decreasing total MLC leaf travel distance and leaf speed.« less
Dynamic simulation of motion effects in IMAT lung SBRT.
Zou, Wei; Yin, Lingshu; Shen, Jiajian; Corradetti, Michael N; Kirk, Maura; Munbodh, Reshma; Fang, Penny; Jabbour, Salma K; Simone, Charles B; Yue, Ning J; Rengan, Ramesh; Teo, Boon-Keng Kevin
2014-11-01
Intensity modulated arc therapy (IMAT) has been widely adopted for Stereotactic Body Radiotherapy (SBRT) for lung cancer. While treatment dose is optimized and calculated on a static Computed Tomography (CT) image, the effect of the interplay between the target and linac multi-leaf collimator (MLC) motion is not well described and may result in deviations between delivered and planned dose. In this study, we investigated the dosimetric consequences of the inter-play effect on target and organs at risk (OAR) by simulating dynamic dose delivery using dynamic CT datasets. Fifteen stage I non-small cell lung cancer (NSCLC) patients with greater than 10 mm tumor motion treated with SBRT in 4 fractions to a dose of 50 Gy were retrospectively analyzed for this study. Each IMAT plan was initially optimized using two arcs. Simulated dynamic delivery was performed by associating the MLC leaf position, gantry angle and delivered beam monitor units (MUs) for each control point with different respiratory phases of the 4D-CT using machine delivery log files containing time stamps of the control points. Dose maps associated with each phase of the 4D-CT dose were calculated in the treatment planning system and accumulated using deformable image registration onto the exhale phase of the 4D-CT. The original IMAT plans were recalculated on the exhale phase of the CT for comparison with the dynamic simulation. The dose coverage of the PTV showed negligible variation between the static and dynamic simulation. There was less than 1.5% difference in PTV V95% and V90%. The average inter-fraction and cumulative dosimetric effects among all the patients were less than 0.5% for PTV V95% and V90% coverage and 0.8 Gy for the OARs. However, in patients where target is close to the organs, large variations were observed on great vessels and bronchus for as much as 4.9 Gy and 7.8 Gy. Limited variation in target dose coverage and OAR constraints were seen for each SBRT fraction as well as over all four fractions. Large dose variations were observed on critical organs in patients where these organs were closer to the target.
Dhabaan, Anees; Elder, Eric; Schreibmann, Eduard; Crocker, Ian; Curran, Walter J; Oyesiku, Nelson M; Shu, Hui-Kuo; Fox, Tim
2010-06-21
The objective was to evaluate the performance of a high-definition multileaf collimator (MLC) of 2.5 mm leaf width (MLC2.5) and compare to standard 5 mm leaf width MLC (MLC5) for the treatment of intracranial lesions using dynamic conformal arcs (DCA) technique with a dedicated radiosurgery linear accelerator. Simulated cases of spherical targets were created to study solely the effect of target volume size on the performance of the two MLC systems independent of target shape complexity. In addition, 43 patients previously treated for intracranial lesions in our institution were retrospectively planned using DCA technique with MLC2.5 and MLC5 systems. The gross tumor volume ranged from 0.07 to 40.57 cm3 with an average volume of 5.9 cm3. All treatment parameters were kept the same for both MLC-based plans. The plan evaluation was performed using figures of merits (FOM) for a rapid and objective assessment on the quality of the two treatment plans for MLC2.5 and MLC5. The prescription isodose surface was selected as the greatest isodose surface covering >or= 95% of the target volume and delivering 95% of the prescription dose to 99% of target volume. A Conformity Index (CI) and conformity distance index (CDI) were used to quantifying the dose conformity to a target volume. To assess normal tissue sparing, a normal tissue difference (NTD) was defined as the difference between the volume of normal tissue receiving a certain dose utilizing MLC5 and the volume receiving the same dose using MLC2.5. The CI and normal tissue sparing for the simulated spherical targets were better with the MLC2.5 as compared to MLC5. For the clinical patients, the CI and CDI results indicated that the MLC2.5 provides better treatment conformity than MLC5 even at large target volumes. The CI's range was 1.15 to 2.44 with a median of 1.59 for MLC2.5 compared to 1.60-2.85 with a median of 1.71 for MLC5. Improved normal tissue sparing was also observed for MLC2.5 over MLC5, with the NTD always positive, indicating improvement, and ranging from 0.1 to 8.3 for normal tissue receiving 50% (NTV50), 70% (NTV70) and 90% (NTV90) of the prescription dose. The MLC2.5 has a dosimetric advantage over the MLC5 in Linac-based radiosurgery using DCA method for intracranial lesions, both in treatment conformity and normal tissue sparing when target shape complexity increases.
Brignone, Maria S.; Lanciotti, Angela; Macioce, Pompeo; Macchia, Gianfranco; Gaetani, Matteo; Aloisi, Francesca; Petrucci, Tamara C.; Ambrosini, Elena
2011-01-01
Megalencephalic leucoencephalopathy with subcortical cysts (MLC) is a rare congenital leucodystrophy caused by mutations in MLC1, a membrane protein of unknown function. MLC1 expression in astrocyte end-feet contacting blood vessels and meninges, along with brain swelling, fluid cysts and myelin vacuolation observed in MLC patients, suggests a possible role for MLC1 in the regulation of fluid and ion homeostasis and cellular volume changes. To identify MLC1 direct interactors and dissect the molecular pathways in which MLC1 is involved, we used NH2-MLC1 domain as a bait to screen a human brain library in a yeast two-hybrid assay. We identified the β1 subunit of the Na,K-ATPase pump as one of the interacting clones and confirmed it by pull-downs, co-fractionation assays and immunofluorescence stainings in human and rat astrocytes in vitro and in brain tissue. By performing ouabain-affinity chromatography on astrocyte and brain extracts, we isolated MLC1 and the whole Na,K-ATPase enzyme in a multiprotein complex that included Kir4.1, syntrophin and dystrobrevin. Because Na,K-ATPase is involved in intracellular osmotic control and volume regulation, we investigated the effect of hypo-osmotic stress on MLC1/Na,K-ATPase relationship in astrocytes. We found that hypo-osmotic conditions increased MLC1 membrane expression and favoured MLC1/Na,K-ATPase-β1 association. Moreover, hypo-osmosis induced astrocyte swelling and the reversible formation of endosome-derived vacuoles, where the two proteins co-localized. These data suggest that through its interaction with Na,K-ATPase, MLC1 is involved in the control of intracellular osmotic conditions and volume regulation in astrocytes, opening new perspectives for understanding the pathological mechanisms of MLC disease. PMID:20926452
Gong, Youling; Wang, Shichao; Zhou, Lin; Liu, Yongmei; Xu, Yong; Lu, You; Bai, Sen; Fu, Yuchuan; Xu, Qingfeng; Jiang, Qingfeng
2010-07-15
To study the impacts of multileaf collimators (MLC) width [standard MLC width of 10 mm (sMLC) and micro-MLC width of 4 mm (mMLC)] in the intensity-modulated radiotherapy (IMRT) planning for the upper thoracic esophageal cancer (UTEC). 10 patients with UTEC were retrospectively planned with the sMLC and the mMLC. The monitor unites (MUs) and dose volume histogram-based parameters [conformity index (CI) and homogeneous index (HI)] were compared between the IMRT plans with sMLC and with mMLC. The IMRT plans with the mMLC were more efficient (average MUs: 703.1 +/- 68.3) than plans with the sMLC (average MUs: 833.4 +/- 73.8) (p < 0.05). Also, compared to plans with the sMLC, the plans with the mMLC showed advantages in dose coverage of the planning gross tumor volume (Pgtv) (CI 0.706 +/- 0.056/HI 1.093 +/- 0.021) and the planning target volume (PTV) (CI 0.707 +/- 0.029/HI 1.315 +/- 0.013) (p < 0.05). In addition, the significant dose sparing in the D5 (3260.3 +/- 374.0 vs 3404.5 +/- 374.4)/gEUD (1815.1 +/- 281.7 vs 1849.2 +/- 297.6) of the spinal cord, the V10 (33.2 +/- 6.5 vs 34.0 +/- 6.7), V20 (16.0 +/- 4.6 vs 16.6 +/- 4.7), MLD (866.2 +/- 174.1 vs 887.9 +/- 172.1) and gEUD (938.6 +/- 175.2 vs 956.8 +/- 171.0) of the lungs were observed in the plans with the mMLC, respectively (p < 0.05). Comparing to the sMLC, the mMLC not only demonstrated higher efficiencies and more optimal target coverage, but also considerably improved the dose sparing of OARs in the IMRT planning for UTEC.
Dubey, Mohit; Brouwers, Eelke; Hamilton, Eline M.C.; Stiedl, Oliver; Bugiani, Marianna; Koch, Henner; Kole, Maarten H.P.; Boschert, Ursula; Wykes, Robert C.; Mansvelder, Huibert D.; van der Knaap, Marjo S.
2018-01-01
Objective Loss of function of the astrocyte‐specific protein MLC1 leads to the childhood‐onset leukodystrophy “megalencephalic leukoencephalopathy with subcortical cysts” (MLC). Studies on isolated cells show a role for MLC1 in astrocyte volume regulation and suggest that disturbed brain ion and water homeostasis is central to the disease. Excitability of neuronal networks is particularly sensitive to ion and water homeostasis. In line with this, reports of seizures and epilepsy in MLC patients exist. However, systematic assessment and mechanistic understanding of seizures in MLC are lacking. Methods We analyzed an MLC patient inventory to study occurrence of seizures in MLC. We used two distinct genetic mouse models of MLC to further study epileptiform activity and seizure threshold through wireless extracellular field potential recordings. Whole‐cell patch‐clamp recordings and K+‐sensitive electrode recordings in mouse brain slices were used to explore the underlying mechanisms of epilepsy in MLC. Results An early onset of seizures is common in MLC. Similarly, in MLC mice, we uncovered spontaneous epileptiform brain activity and a lowered threshold for induced seizures. At the cellular level, we found that although passive and active properties of individual pyramidal neurons are unchanged, extracellular K+ dynamics and neuronal network activity are abnormal in MLC mice. Interpretation Disturbed astrocyte regulation of ion and water homeostasis in MLC causes hyperexcitability of neuronal networks and seizures. These findings suggest a role for defective astrocyte volume regulation in epilepsy. Ann Neurol 2018;83:636–649 PMID:29466841
Storage of sparse files using parallel log-structured file system
DOE Office of Scientific and Technical Information (OSTI.GOV)
Bent, John M.; Faibish, Sorin; Grider, Gary
A sparse file is stored without holes by storing a data portion of the sparse file using a parallel log-structured file system; and generating an index entry for the data portion, the index entry comprising a logical offset, physical offset and length of the data portion. The holes can be restored to the sparse file upon a reading of the sparse file. The data portion can be stored at a logical end of the sparse file. Additional storage efficiency can optionally be achieved by (i) detecting a write pattern for a plurality of the data portions and generating a singlemore » patterned index entry for the plurality of the patterned data portions; and/or (ii) storing the patterned index entries for a plurality of the sparse files in a single directory, wherein each entry in the single directory comprises an identifier of a corresponding sparse file.« less
Pickering, Bradley S; Lopilato, Jane E; Smith, Daniel R; Watnick, Paula I
2014-07-01
The phosphoenol phosphotransferase system (PTS) is a multicomponent signal transduction cascade that regulates diverse aspects of bacterial cellular physiology in response to the availability of high-energy sugars in the environment. Many PTS components are repressed at the transcriptional level when the substrates they transport are not available. In Escherichia coli, the transcription factor Mlc (for makes large colonies) represses transcription of the genes encoding enzyme I (EI), histidine protein (HPr), and the glucose-specific enzyme IIBC (EIIBC(Glc)) in defined media that lack PTS substrates. When glucose is present, the unphosphorylated form of EIIBC(Glc) sequesters Mlc to the cell membrane, preventing its interaction with DNA. Very little is known about Vibrio cholerae Mlc. We found that V. cholerae Mlc activates biofilm formation in LB broth but not in defined medium supplemented with either pyruvate or glucose. Therefore, we questioned whether V. cholerae Mlc functions differently than E. coli Mlc. Here we have shown that, like E. coli Mlc, V. cholerae Mlc represses transcription of PTS components in both defined medium and LB broth and that E. coli Mlc is able to rescue the biofilm defect of a V. cholerae Δmlc mutant. Furthermore, we provide evidence that Mlc indirectly activates transcription of the vps genes by repressing expression of EI. Because activation of the vps genes by Mlc occurs under only a subset of the conditions in which repression of PTS components is observed, we conclude that additional inputs present in LB broth are required for activation of vps gene transcription by Mlc. Copyright © 2014, American Society for Microbiology. All Rights Reserved.
Technology assessment of multileaf collimation: a North American users survey.
Klein, E E; Tepper, J; Sontag, M; Franklin, M; Ling, C; Kubo, D
1999-06-01
The American Association of Physicists in Medicine (AAPM) initiated an Assessment of Technology Subcommittee (ATS) to help the radiotherapy community evaluate emerging technologies. The ATS decided to first address multileaf collimation (MLC) by means of a North American users survey. The survey attempted to address issues such as MLC utility, efficacy, cost-effectiveness, and customer satisfaction. The survey was designed with 38 questions, with cross-tabulation set up to decipher a particular clinic's perception of MLC. The surveys were coded according to MLC types, which were narrowed to four: Elekta, Siemens, Varian 52-leaf, and Varian 80-leaf. A 40% return rate was desired. A 44% (108 of 250) return was achieved. On an MLC machine, 76.5% of photon patients are being treated with MLC. The main reasons for not using MLC were stair stepping, field size limitation, and physician objection. The most common sites in which MLC is being used are lung, pelvis, and prostate. The least used sites are head & neck and mantle fields. Of the facilities, 31% claimed an increase in number of patients being treated since MLC was installed, and 44% claimed an increase in the number of fields. Though the staffing for block cutting has decreased, therapist staffing has not. However, 91% of the facilities claimed a decreased workload for the therapists, despite the increase in daily treated patients and fields. Of the facilities that justified MLC purchase for more daily patients, 63% are actually treating more patients. Only 26% of the facilities that justified an MLC purchase for intensity-modulated radiotherapy (IMRT) are currently using it for that purpose. The satisfaction rating (1 = low to 5 = high) for department groups averaged 4.0. Therapists ranked MLC as 4.6. Our survey shows that most users have successfully introduced MLC into the clinic as a block replacement. Most have found MLC to be cost-effective and efficient. The use of MLC for IMRT has progressed slower, but users anticipate escalated use.
SU-E-T-178: Clinical Feasibility of Multi-Leaf Collimator Based Dynamic Wedge
DOE Office of Scientific and Technical Information (OSTI.GOV)
Jeong, C; Kwak, J; Ahn, S
2015-06-15
Purpose: A multi-leaf collimator (MLC) based dynamic wedge (MDW), which provide similar dose profile of physical wedge (PW) along x-jaw direction while significant monitor unit (MU) reduction, was developed and investigated for clinical use. Methods: A novel technique was used to create the wedge profile using MLC. A modification was applied to the DICOM-RT format file of the plan made with the PW to replace PW with MDW. The Varian enhanced dynamic wedge profile was used to produce MLC sequence, while the MU of the wedged field was recalculated using PW factor and fluence map. The profiles for all possiblemore » MDWs to substitute PWs were verified in 6/15 MV x-ray irradiations. New plans with MDWs were compared with the original plans in 5 rectal, 5 RT breast and 5 liver cases. Results: The wedge profile of the MDW fields were well matched with those of PWs inside the fields while less scatter than PW out of the fields. For plan comparisons of the clinical cases no significant dose discrepancy was observed between MDW plan and PW’s with the dose volume histograms. The maximum and mean doses in PTVs are agreed within 1.0%. The Result of OARs of MDW plans are slightly improved in the maximum doses (3.22 ∼ 150.4 cGy) and the mean doses (17.18 ∼ 85.52 cGy) on average for all cases while the prescribed doses are 45 Gy for rectal cases, 40 or 45 Gy for liver cases and 50 Gy for breast cases. The MUs of the fields which replace PW with MDW are reduced to 68% of those of PW. Conclusion: We developed a novel dynamic wedge technique with MLC that shows clinical advantage compared to PW.« less
The potential impact of multidimesional geriatric assessment in the social security system.
Corbi, Graziamaria; Ambrosino, Immacolata; Massari, Marco; De Lucia, Onofrio; Simplicio, Sirio; Dragone, Michele; Paolisso, Giuseppe; Piccioni, Massimo; Ferrara, Nicola; Campobasso, Carlo Pietro
2018-01-12
To evaluate the efficacy of multidimensional geriatric assessment (MGA/CGA) in patients over 65 years old in predicting the release of the accompaniment allowance (AA) indemnity by a Local Medico-Legal Committee (MLC-NHS) and by the National Institute of Social Security Committee (MLC-INPS). In a longitudinal observational study, 200 Italian elder citizens requesting AA were first evaluated by MLC-NHS and later by MLC-INPS. Only MLC-INPS performed a MGA/CGA (including SPMSQ, Barthel Index, GDS-SF, and CIRS). This report was written according to the STROBE guidelines. The data analysis was performed on January 2016. The evaluation by the MLC-NHS and by the MLC-INPS was in agreement in 66% of cases. In the 28%, the AA benefit was recognized by the MLC-NHS, but not by the MLC-INPS. By the multivariate analysis, the best predictors of the AA release, by the MLC-NHS, were represented by gender and the Barthel Index score. The presence of carcinoma, the Barthel Index score, and the SPMQ score were the best predictors for the AA release by MLC-INPS. MGA/CGA could be useful in saving financial resources reducing the risk of incorrect indemnity release. It can improve the accuracy of the impairment assessment in social security system.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Park, So-Yeon; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul 110-744; Biomedical Research Institute, Seoul National University College of Medicine, Seoul 110-744
Purpose: Texture analysis on fluence maps was performed to evaluate the degree of modulation for volumetric modulated arc therapy (VMAT) plans. Methods: A total of six textural features including angular second moment, inverse difference moment, contrast, variance, correlation, and entropy were calculated for fluence maps generated from 20 prostate and 20 head and neck VMAT plans. For each of the textural features, particular displacement distances (d) of 1, 5, and 10 were adopted. To investigate the deliverability of each VMAT plan, gamma passing rates of pretreatment quality assurance, and differences in modulating parameters such as multileaf collimator (MLC) positions, gantrymore » angles, and monitor units at each control point between VMAT plans and dynamic log files registered by the Linac control system during delivery were acquired. Furthermore, differences between the original VMAT plan and the plan reconstructed from the dynamic log files were also investigated. To test the performance of the textural features as indicators for the modulation degree of VMAT plans, Spearman’s rank correlation coefficients (r{sub s}) with the plan deliverability were calculated. For comparison purposes, conventional modulation indices for VMAT including the modulation complexity score for VMAT, leaf travel modulation complexity score, and modulation index supporting station parameter optimized radiation therapy (MI{sub SPORT}) were calculated, and their correlations were analyzed in the same way. Results: There was no particular textural feature which always showed superior correlations with every type of plan deliverability. Considering the results comprehensively, contrast (d = 1) and variance (d = 1) generally showed considerable correlations with every type of plan deliverability. These textural features always showed higher correlations to the plan deliverability than did the conventional modulation indices, except in the case of modulating parameter differences. The r{sub s} values of contrast to the global gamma passing rates with criteria of 2%/2 mm, 2%/1 mm, and 1%/2 mm were 0.536, 0.473, and 0.718, respectively. The respective values for variance were 0.551, 0.481, and 0.688. In the case of local gamma passing rates, the r{sub s} values of contrast were 0.547, 0.578, and 0.620, respectively, and those of variance were 0.519, 0.527, and 0.569. All of the r{sub s} values in those cases were statistically significant (p < 0.003). In the cases of global and local gamma passing rates, MI{sub SPORT} showed the highest correlations among the conventional modulation indices. For global passing rates, r{sub s} values of MI{sub SPORT} were −0.420, −0.330, and −0.632, respectively, and those for local passing rates were −0.455, −0.490 and −0.502. The values of r{sub s} of contrast, variance, and MI{sub SPORT} with the MLC errors were −0.863, −0.828, and 0.795, respectively, all with statistical significances (p < 0.001). The correlations with statistical significances between variance and dose-volumetric differences were observed more frequently than the others. Conclusions: The contrast (d = 1) and variance (d = 1) calculated from fluence maps of VMAT plans showed considerable correlations with the plan deliverability, indicating their potential use as indicators for assessing the degree of modulation of VMAT plans. Both contrast and variance consistently showed better performance than the conventional modulation indices for VMAT.« less
A method for photon beam Monte Carlo multileaf collimator particle transport
NASA Astrophysics Data System (ADS)
Siebers, Jeffrey V.; Keall, Paul J.; Kim, Jong Oh; Mohan, Radhe
2002-09-01
Monte Carlo (MC) algorithms are recognized as the most accurate methodology for patient dose assessment. For intensity-modulated radiation therapy (IMRT) delivered with dynamic multileaf collimators (DMLCs), accurate dose calculation, even with MC, is challenging. Accurate IMRT MC dose calculations require inclusion of the moving MLC in the MC simulation. Due to its complex geometry, full transport through the MLC can be time consuming. The aim of this work was to develop an MLC model for photon beam MC IMRT dose computations. The basis of the MC MLC model is that the complex MLC geometry can be separated into simple geometric regions, each of which readily lends itself to simplified radiation transport. For photons, only attenuation and first Compton scatter interactions are considered. The amount of attenuation material an individual particle encounters while traversing the entire MLC is determined by adding the individual amounts from each of the simplified geometric regions. Compton scatter is sampled based upon the total thickness traversed. Pair production and electron interactions (scattering and bremsstrahlung) within the MLC are ignored. The MLC model was tested for 6 MV and 18 MV photon beams by comparing it with measurements and MC simulations that incorporate the full physics and geometry for fields blocked by the MLC and with measurements for fields with the maximum possible tongue-and-groove and tongue-or-groove effects, for static test cases and for sliding windows of various widths. The MLC model predicts the field size dependence of the MLC leakage radiation within 0.1% of the open-field dose. The entrance dose and beam hardening behind a closed MLC are predicted within +/-1% or 1 mm. Dose undulations due to differences in inter- and intra-leaf leakage are also correctly predicted. The MC MLC model predicts leaf-edge tongue-and-groove dose effect within +/-1% or 1 mm for 95% of the points compared at 6 MV and 88% of the points compared at 18 MV. The dose through a static leaf tip is also predicted generally within +/-1% or 1 mm. Tests with sliding windows of various widths confirm the accuracy of the MLC model for dynamic delivery and indicate that accounting for a slight leaf position error (0.008 cm for our MLC) will improve the accuracy of the model. The MLC model developed is applicable to both dynamic MLC and segmental MLC IMRT beam delivery and will be useful for patient IMRT dose calculations, pre-treatment verification of IMRT delivery and IMRT portal dose transmission dosimetry.
46 CFR 97.35-3 - Logbooks and records.
Code of Federal Regulations, 2010 CFR
2010-10-01
... voyage is completed, the master or person in charge shall file the logbook with the Officer in Charge.... Such logs or records are not filed with the Officer in Charge, Marine Inspection, but must be kept... logs for the period of validity of the vessel's certificate of inspection. [CGD 95-027, 61 FR 26007...
46 CFR 97.35-3 - Logbooks and records.
Code of Federal Regulations, 2011 CFR
2011-10-01
... voyage is completed, the master or person in charge shall file the logbook with the Officer in Charge.... Such logs or records are not filed with the Officer in Charge, Marine Inspection, but must be kept... logs for the period of validity of the vessel's certificate of inspection. [CGD 95-027, 61 FR 26007...
DOE Office of Scientific and Technical Information (OSTI.GOV)
J Zwan, B; Central Coast Cancer Centre, Gosford, NSW; Colvill, E
2016-06-15
Purpose: The added complexity of the real-time adaptive multi-leaf collimator (MLC) tracking increases the likelihood of undetected MLC delivery errors. In this work we develop and test a system for real-time delivery verification and error detection for MLC tracking radiotherapy using an electronic portal imaging device (EPID). Methods: The delivery verification system relies on acquisition and real-time analysis of transit EPID image frames acquired at 8.41 fps. In-house software was developed to extract the MLC positions from each image frame. Three comparison metrics were used to verify the MLC positions in real-time: (1) field size, (2) field location and, (3)more » field shape. The delivery verification system was tested for 8 VMAT MLC tracking deliveries (4 prostate and 4 lung) where real patient target motion was reproduced using a Hexamotion motion stage and a Calypso system. Sensitivity and detection delay was quantified for various types of MLC and system errors. Results: For both the prostate and lung test deliveries the MLC-defined field size was measured with an accuracy of 1.25 cm{sup 2} (1 SD). The field location was measured with an accuracy of 0.6 mm and 0.8 mm (1 SD) for lung and prostate respectively. Field location errors (i.e. tracking in wrong direction) with a magnitude of 3 mm were detected within 0.4 s of occurrence in the X direction and 0.8 s in the Y direction. Systematic MLC gap errors were detected as small as 3 mm. The method was not found to be sensitive to random MLC errors and individual MLC calibration errors up to 5 mm. Conclusion: EPID imaging may be used for independent real-time verification of MLC trajectories during MLC tracking deliveries. Thresholds have been determined for error detection and the system has been shown to be sensitive to a range of delivery errors.« less
Micellar liquid chromatography
NASA Astrophysics Data System (ADS)
Basova, Elena M.; Ivanov, Vadim M.; Shpigun, Oleg A.
1999-12-01
Background and possibilities of practical applications of micellar liquid chromatography (MLC) are considered. Various retention models in MLC, the effects of the nature and concentration of surfactants and organic modifiers, pH, temperature and ionic strength on the MLC efficiency and selectivity are discussed. The advantages and limitations of MLC are demonstrated. The performance of MLC is critically evaluated in relationship to the reversed-phase HPLC and ion-pair chromatography. The potential of application of MLC for the analysis of pharmaceuticals including that in biological fluids and separation of inorganic anions, transition metal cations, metal chelates and heteropoly compounds is described. The bibliography includes 146 references.
Lee, Gyeongmin; Kim, Yoonbin; Kim, Hoikyung; Beuchat, Larry R; Ryu, Jee-Hoon
2018-01-16
The aim of this study was to evaluate the antimicrobial activities of gaseous essential oils (EO gases) against Listeria monocytogenes on the surfaces of a laboratory medium and radish sprouts. We determined the minimal inhibitory concentration (MIC) and minimal lethal concentration (MLC) values of EO gases from eight EOs extracted from basil leaves, carrot seed, cinnamon bark, cinnamon leaves, clove flower buds, oregano leaves, thyme flowers (linalool), and thyme leaves (thymol) against L. monocytogenes on a nutrient agar supplemented with 1% glucose and 0.025% bromocresol purple (NGBA). Oregano, thyme thymol, and cinnamon bark EO gases showed the strongest antilisterial activities (MIC and MLC, 78.1μL/L). We also investigated the inhibitory and lethal activities of these gases against L. monocytogenes on the surface of radish sprouts. The number of L. monocytogenes after exposure to EO gases at ≥156μL/L was significantly (P≤0.05) lower than that of untreated L. monocytogenes. For example, the initial number of L. monocytogenes on the surface of radish sprouts (ca. 6.3logCFU/g) decreased by 1.4logCFU/g within 24h at 30°C and 43% relative humidity (RH) without EO gas treatment, whereas the number of L. monocytogenes after exposure to oregano, thyme thymol, and cinnamon bark EO gases at 156μL/L decreased by 2.1, 2.1, and 1.8logCFU/g, respectively, after 24h. Although EO gases exerted greater lethal activities at higher concentrations (312 and 625μL/L), L. monocytogenes on the surface of radish sprouts was not completely inactivated. The number of L. monocytogenes on sprouts treated with oregano, thyme thymol, and cinnamon bark EO gases at 625μL/L decreased by 2.7-3.0logCFU/g after 24h at 30°C and 43% RH. Results indicate that EO gases that showed antilisterial activities on a laboratory medium also exhibited reduced lethal activity on the surface of radish sprouts. These findings will be useful when developing strategies to inactivate L. monocytogenes and possibly other foodborne pathogens on sprouts and perhaps other foods using EO gases. Copyright © 2017 Elsevier B.V. All rights reserved.
Electromagnetic guided couch and multileaf collimator tracking on a TrueBeam accelerator
DOE Office of Scientific and Technical Information (OSTI.GOV)
Hansen, Rune; Ravkilde, Thomas; Worm, Esben Schjødt
2016-05-15
Purpose: Couch and MLC tracking are two promising methods for real-time motion compensation during radiation therapy. So far, couch and MLC tracking experiments have mainly been performed by different research groups, and no direct comparison of couch and MLC tracking of volumetric modulated arc therapy (VMAT) plans has been published. The Varian TrueBeam 2.0 accelerator includes a prototype tracking system with selectable couch or MLC compensation. This study provides a direct comparison of the two tracking types with an otherwise identical setup. Methods: Several experiments were performed to characterize the geometric and dosimetric performance of electromagnetic guided couch and MLCmore » tracking on a TrueBeam accelerator equipped with a Millennium MLC. The tracking system latency was determined without motion prediction as the time lag between sinusoidal target motion and the compensating motion of the couch or MLC as recorded by continuous MV portal imaging. The geometric and dosimetric tracking accuracies were measured in tracking experiments with motion phantoms that reproduced four prostate and four lung tumor trajectories. The geometric tracking error in beam’s eye view was determined as the distance between an embedded gold marker and a circular MLC aperture in continuous MV images. The dosimetric tracking error was quantified as the measured 2%/2 mm gamma failure rate of a low and a high modulation VMAT plan delivered with the eight motion trajectories using a static dose distribution as reference. Results: The MLC tracking latency was approximately 146 ms for all sinusoidal period lengths while the couch tracking latency increased from 187 to 246 ms with decreasing period length due to limitations in the couch acceleration. The mean root-mean-square geometric error was 0.80 mm (couch tracking), 0.52 mm (MLC tracking), and 2.75 mm (no tracking) parallel to the MLC leaves and 0.66 mm (couch), 1.14 mm (MLC), and 2.41 mm (no tracking) perpendicular to the leaves. The motion-induced gamma failure rate was in mean 0.1% (couch tracking), 8.1% (MLC tracking), and 30.4% (no tracking) for prostate motion and 2.9% (couch), 2.4% (MLC), and 41.2% (no tracking) for lung tumor motion. The residual tracking errors were mainly caused by inadequate adaptation to fast lung tumor motion for couch tracking and to prostate motion perpendicular to the MLC leaves for MLC tracking. Conclusions: Couch and MLC tracking markedly improved the geometric and dosimetric accuracies of VMAT delivery. However, the two tracking types have different strengths and weaknesses. While couch tracking can correct perfectly for slowly moving targets such as the prostate, MLC tracking may have considerably larger dose errors for persistent target shift perpendicular to the MLC leaves. Advantages of MLC tracking include faster dynamics with better adaptation to fast moving targets, the avoidance of moving the patient, and the potential to track target rotations and deformations.« less
Crangle, Robert D.
2007-01-01
Introduction The U.S. Geological Survey (USGS) uses geophysical wireline well logs for a variety of purposes, including stratigraphic correlation (Hettinger, 2001, Ryder, 2002), petroleum reservoir analyses (Nelson and Bird, 2005), aquifer studies (Balch, 1988), and synthetic seismic profiles (Kulander and Ryder, 2005). Commonly, well logs are easier to visualize, manipulate, and interpret when available in a digital format. In recent geologic cross sections E-E' and D-D', constructed through the central Appalachian basin (Ryder, Swezey, and others, in press; Ryder, Crangle, and others, in press), gamma ray well log traces and lithologic logs were used to correlate key stratigraphic intervals (Fig. 1). The stratigraphy and structure of the cross sections are illustrated through the use of graphical software applications (e.g., Adobe Illustrator). The gamma ray traces were digitized in Neuralog (proprietary software) from paper well logs and converted to a Log ASCII Standard (LAS) format. Once converted, the LAS files were transformed to images through an LAS-reader application (e.g., GeoGraphix Prizm) and then overlain in positions adjacent to well locations, used for stratigraphic control, on each cross section. This report summarizes the procedures used to convert paper logs to a digital LAS format using a third-party software application, Neuralog. Included in this report are LAS files for sixteen wells used in geologic cross section E-E' (Table 1) and thirteen wells used in geologic cross section D-D' (Table 2).
Is mini-laparoscopic cholecystectomy any better than the gold standard?: A comparative study.
Shaikh, Haris R; Abbas, Asad; Aleem, Salik; Lakhani, Miqdad R
2017-01-01
Mini-laparoscopic cholecystectomy (MLC) has widened the horizons of modern laparoscopic surgery. Standard four port laparoscopic cholecystectomy (SLC), which has long been established as the "Gold Standard" for gall bladder diseases, is under reconsideration following the advent of further minimally-invasive procedures including MLC. Our study aims to provide a comparison between MLC and SLC and assesses whether MLC has any added benefits. Patients with symptomatic gall bladder disease undergoing MLC or SLC during the 2.5-month period were included in the study. Thirty-two patients underwent MLC while SLC was performed on 40 patients by the same surgeon. Data was collected prospectively and analysed retrospectively using a predesigned questionnaire. In our study, both the groups had similar age, body mass index (BMI) and gender distribution. No cases of MLC required insertion of additional ports. The mean operative time for MLC was 38.2 min (33-61 min), which is longer than SLC; but it was not statistically significant. There was no significant difference in mean operative blood loss, postoperative pain, analgesia requirement and mobilization. Patients who underwent MLC were able to return to normal activity earlier than patients undergoing SLC (P < 0.01). Our experience suggests that MLC can safely be used as an alternative to SLC. Compared to SLC, it has the added benefit of an early return to work along with excellent cosmetic results. Further large scale trials are required to prove any additional benefit of MLC.
Is mini-laparoscopic cholecystectomy any better than the gold standard?: A comparative study
Shaikh, Haris R.; Abbas, Asad; Aleem, Salik; Lakhani, Miqdad R.
2017-01-01
BACKGROUND: Mini-laparoscopic cholecystectomy (MLC) has widened the horizons of modern laparoscopic surgery. Standard four port laparoscopic cholecystectomy (SLC), which has long been established as the “Gold Standard” for gall bladder diseases, is under reconsideration following the advent of further minimally-invasive procedures including MLC. Our study aims to provide a comparison between MLC and SLC and assesses whether MLC has any added benefits. MATERIALS AND METHODS: Patients with symptomatic gall bladder disease undergoing MLC or SLC during the 2.5-month period were included in the study. Thirty-two patients underwent MLC while SLC was performed on 40 patients by the same surgeon. Data was collected prospectively and analysed retrospectively using a predesigned questionnaire. RESULTS: In our study, both the groups had similar age, body mass index (BMI) and gender distribution. No cases of MLC required insertion of additional ports. The mean operative time for MLC was 38.2 min (33-61 min), which is longer than SLC; but it was not statistically significant. There was no significant difference in mean operative blood loss, postoperative pain, analgesia requirement and mobilization. Patients who underwent MLC were able to return to normal activity earlier than patients undergoing SLC (P < 0.01). CONCLUSION: Our experience suggests that MLC can safely be used as an alternative to SLC. Compared to SLC, it has the added benefit of an early return to work along with excellent cosmetic results. Further large scale trials are required to prove any additional benefit of MLC. PMID:27251827
Jang, Si Young; Lalonde, Ron; Ozhasoglu, Cihat; Burton, Steven; Heron, Dwight; Huq, M Saiful
2016-09-08
We performed an evaluation of the CyberKnife InCise MLC by comparing plan qualities for single and multiple brain lesions generated using the first version of InCise MLC, fixed cone, and Iris collimators. We also investigated differences in delivery efficiency among the three collimators. Twenty-four patients with single or multiple brain mets treated previously in our clinic on a CyberKnife M6 using cone/Iris collimators were selected for this study. Treatment plans were generated for all lesions using the InCise MLC. Number of monitor units, delivery time, target coverage, conformity index, and dose falloff were compared between MLC- and clinical cone/Iris-based plans. Statistical analysis was performed using the non-parametric Wilcoxon-Mann-Whitney signed-rank test. The planning accuracy of the MLC-based plans was validated using chamber and film measurements. The InCise MLC-based plans achieved mean dose and target coverage comparable to the cone/Iris-based plans. Although the conformity indices of the MLC-based plans were slightly higher than those of the cone/Iris-based plans, beam delivery time for the MLC-based plans was shorter by 30% ~ 40%. For smaller targets or cases with OARs located close to or abutting target volumes, MLC-based plans provided inferior dose conformity compared to cone/Iris-based plans. The QA results of MLC-based plans were within 5% absolute dose difference with over 90% gamma passing rate using 2%/2 mm gamma criteria. The first version of InCise MLC could be a useful delivery modality, especially for clinical situations for which delivery time is a limiting factor or for multitarget cases. © 2016 The Authors.
SU-F-T-463: Light-Field Based Dynalog Verification
DOE Office of Scientific and Technical Information (OSTI.GOV)
Atwal, P; Ramaseshan, R
2016-06-15
Purpose: To independently verify leaf positions in so-called dynalog files for a Varian iX linac with a Millennium 120 MLC. This verification provides a measure of confidence that the files can be used directly as part of a more extensive intensity modulated radiation therapy / volumetric modulated arc therapy QA program. Methods: Initial testing used white paper placed at the collimator plane and a standard hand-held digital camera to image the light and shadow of a static MLC field through the paper. Known markings on the paper allow for image calibration. Noise reduction was attempted with removal of ‘inherent noise’more » from an open-field light image through the paper, but the method was found to be inconsequential. This is likely because the environment could not be controlled to the precision required for the sort of reproducible characterization of the quantum noise needed in order to meaningfully characterize and account for it. A multi-scale iterative edge detection algorithm was used for localizing the leaf ends. These were compared with the planned locations from the treatment console. Results: With a very basic setup, the image of the central bank A leaves 15–45, which are arguably the most important for beam modulation, differed from the planned location by [0.38±0.28] mm. Similarly, for bank B leaves 15–45 had a difference of [0.42±0.28] mm Conclusion: It should be possible to determine leaf position accurately with not much more than a modern hand-held camera and some software. This means we can have a periodic and independent verification of the dynalog file information. This is indicated by the precision already achieved using a basic setup and analysis methodology. Currently, work is being done to reduce imaging and setup errors, which will bring the leaf position error down further, and allow meaningful analysis over the full range of leaves.« less
Multileaf collimator characteristics and reliability requirements for IMRT Elekta system.
Liu, Chihray; Simon, Thomas A; Fox, Christopher; Li, Jonathan; Palta, Jatinder R
2008-01-01
Understanding the characteristics of a multileaf collimator (MLC) system, modeling MLC in a treatment planning system, and maintaining the mechanical accuracy of the linear accelerator gantry head system are important factors in the safe implementation of an intensity-modulated radiotherapy program. We review the characteristics of an Elekta MLC system, discuss the necessary MLC modeling parameters for a treatment planning system, and provide a novel method to establish an MLC leaf position quality assurance program. To perform quality assurance on 40 pairs of individual MLC leaves is a time-consuming and difficult task. In this report, an effective routine MLC quality assurance method based on the field edge of a backup jaw as referenced in conjunction with a diode array as a radiation detector system is discussed. The sensitivity of this test for determining the relative leaf positions was observed to be better than 0.1 mm. The Elekta MLC leaf position accuracy measured with this system has been better than 0.3 mm.
SU-F-T-530: Characterization of a 60-Leaf Motorized MLC Designed for Cobalt-60 Units
DOE Office of Scientific and Technical Information (OSTI.GOV)
Xu, L; Smith, L; Ciresianu, A
2016-06-15
Purpose: In a continuing effort to improve conformal radiation therapy with Cobalt-60 units, a 60-leaf MLC was designed, manufactured, and released to market. This work describes the physics measurements taken to characterize the clinical performance of this MLC. Methods: A 60 leaf MLC was custom designed with tungsten leaves of 4.5 cm height, single focused, achieving field size of 30×30 cm^2 when mounted on a 100cm SAD Cobalt-60 unit. Leakage and output factor measurements were performed using a single ion chamber in a solid water phantom. Penumbra and surface dose were measured using scanning chambers and diodes in a watermore » phantom. Radiation-light coincidence measurements were performed using radiographic films. Results: With MLC mounted, measured penumbras at all depths are smaller than with jaws only. Surface doses were not significantly affected by the presence of MLC, and remained below values recommended by regulatory bodies. Light-radiation coincidences were found to be better than 3 mm for all field sizes. Leakage through the MLC was found to be strongly dependent on field size, increasing from 1.0 % for a 10×10 cm field to 2.0% for a 30×30 cm field. Such results meet the requirements of IEC 60601-2-11. The MLC was found to have significant influence on the output factor, when field size defined by MLC is significantly smaller than field size defined by jaws. Such effect is also observed on linear accelerators, but it is more pronounced on Cobalt-60 units. A 10×10 “diamond” MLC shape inside a 14×14 cm jaw showed output factor that is 5.7% higher than 10×10 cm field defined by matching MLC and jaws. Conclusion: The MLC offers clinically acceptable performance in penumbra, surface dose, and light-radiation coincidence. Several units of this MLC have recently been installed and used clinically. Validation of Cobalt-60 based IMRT with this MLC is ongoing. The authors are employees of Best Theratrnics Ltd.« less
NASA Astrophysics Data System (ADS)
Hoegg-Beiler, Maja B.; Sirisi, Sònia; Orozco, Ian J.; Ferrer, Isidre; Hohensee, Svea; Auberson, Muriel; Gödde, Kathrin; Vilches, Clara; de Heredia, Miguel López; Nunes, Virginia; Estévez, Raúl; Jentsch, Thomas J.
2014-03-01
Defects in the astrocytic membrane protein MLC1, the adhesion molecule GlialCAM or the chloride channel ClC-2 underlie human leukoencephalopathies. Whereas GlialCAM binds ClC-2 and MLC1, and modifies ClC-2 currents in vitro, no functional connections between MLC1 and ClC-2 are known. Here we investigate this by generating loss-of-function Glialcam and Mlc1 mouse models manifesting myelin vacuolization. We find that ClC-2 is unnecessary for MLC1 and GlialCAM localization in brain, whereas GlialCAM is important for targeting MLC1 and ClC-2 to specialized glial domains in vivo and for modifying ClC-2’s biophysical properties specifically in oligodendrocytes (OLs), the cells chiefly affected by vacuolization. Unexpectedly, MLC1 is crucial for proper localization of GlialCAM and ClC-2, and for changing ClC-2 currents. Our data unmask an unforeseen functional relationship between MLC1 and ClC-2 in vivo, which is probably mediated by GlialCAM, and suggest that ClC-2 participates in the pathogenesis of megalencephalic leukoencephalopathy with subcortical cysts.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Mullins, J; Asiev, K; DeBlois, F
2014-08-15
The TrueBeam linear accelerator platform has a developer's mode which permits the user dynamic control over many of the machine's mechanical and radiation systems. Using this research tool, synchronous couch and gantry motion can be programmed to simulate isocentric treatment with a shortened SAD, with benefits such as smaller projected MLC leaf widths and an increased dose rate. In this work, water tank measurements were used to commission a virtual linear accelerator with an 85 cm SAD in Eclipse, from which several arc-based radiotherapy treatments were generated, including an inverse optimized VMAT delivery. For each plan, the pertinent treatment deliverymore » information was extracted from control points specified in the Eclipse-exported DICOM files using the pydicom package in Python, allowing construction of an XML control file. The dimensions of the jaws and MLC positions, defined for an 85 cm SAD in Eclipse, were scaled for delivery on a conventional SAD linear accelerator, and translational couch motion was added as a function of gantry angle to simulate delivery at 85 cm SAD. Ionization chamber and Gafchromic film measurements were used to compare the radiation delivery to dose calculations in Eclipse. With the exception of the VMAT delivery, ionization chamber measurements agreed within 3.3% of the Eclipse calculations. For the VMAT delivery, the ionization chamber was located in an inhomogeneous region, but gamma evaluation of the Gafchromic film plane resulted in a 94.5% passing rate using criteria of 3 mm/3%. The results indicate that Eclipse calculation infrastructure can be used.« less
20 CFR 401.85 - Exempt systems.
Code of Federal Regulations, 2011 CFR
2011-04-01
... subsection (k)(2) of the Privacy Act: (A) The General Criminal Investigation Files, SSA; (B) The Criminal Investigations File, SSA; and, (C) The Program Integrity Case Files, SSA. (D) Civil and Administrative Investigative Files of the Inspector General, SSA/OIG. (E) Complaint Files and Log. SSA/OGC. (iii) Pursuant to...
47 CFR 76.1706 - Signal leakage logs and repair records.
Code of Federal Regulations, 2010 CFR
2010-10-01
... the probable cause of the leakage. The log shall be kept on file for a period of two years and shall... 47 Telecommunication 4 2010-10-01 2010-10-01 false Signal leakage logs and repair records. 76.1706... leakage logs and repair records. Cable operators shall maintain a log showing the date and location of...
47 CFR 76.1706 - Signal leakage logs and repair records.
Code of Federal Regulations, 2011 CFR
2011-10-01
... the probable cause of the leakage. The log shall be kept on file for a period of two years and shall... 47 Telecommunication 4 2011-10-01 2011-10-01 false Signal leakage logs and repair records. 76.1706... leakage logs and repair records. Cable operators shall maintain a log showing the date and location of...
Inhibiting Myosin Light Chain Kinase Induces Apoptosis In Vitro and In Vivo
Fazal, Fabeha; Gu, Lianzhi; Ihnatovych, Ivanna; Han, YooJeong; Hu, WenYang; Antic, Nenad; Carreira, Fernando; Blomquist, James F.; Hope, Thomas J.; Ucker, David S.; de Lanerolle, Primal
2005-01-01
Previous short-term studies have correlated an increase in the phosphorylation of the 20-kDa light chain of myosin II (MLC20) with blebbing in apoptotic cells. We have found that this increase in MLC20 phosphorylation is rapidly followed by MLC20 dephosphorylation when cells are stimulated with various apoptotic agents. MLC20 dephosphorylation is not a consequence of apoptosis because MLC20 dephosphorylation precedes caspase activation when cells are stimulated with a proapoptotic agent or when myosin light chain kinase (MLCK) is inhibited pharmacologically or by microinjecting an inhibitory antibody to MLCK. Moreover, blocking caspase activation increased cell survival when MLCK is inhibited or when cells are treated with tumor necrosis factor alpha. Depolymerizing actin filaments or detaching cells, processes that destabilize the cytoskeleton, or inhibiting myosin ATPase activity also resulted in MLC20 dephosphorylation and cell death. In vivo experiments showed that inhibiting MLCK increased the number of apoptotic cells and retarded the growth of mammary cancer cells in mice. Thus, MLC20 dephosphorylation occurs during physiological cell death and prolonged MLC20 dephosphorylation can trigger apoptosis. PMID:15988034
Feng, Zhonghui; Okada, Satoshi; Cai, Guoping; Zhou, Bing; Bi, Erfei
2015-01-01
MLC1 is a haploinsufficient gene encoding the essential light chain for Myo1, the sole myosin‑II heavy chain in the budding yeast Saccharomyces cerevisiae. Mlc1 defines an essential hub that coordinates actomyosin ring function, membrane trafficking, and septum formation during cytokinesis by binding to IQGAP, myosin‑II, and myosin‑V. However, the mechanism of how Mlc1 is targeted to the division site during the cell cycle remains unsolved. By constructing a GFP‑tagged MLC1 under its own promoter control and using quantitative live‑cell imaging coupled with yeast mutants, we found that septin ring and actin filaments mediate the targeting of Mlc1 to the division site before and during cytokinesis, respectively. Both mechanisms contribute to and are collectively required for the accumulation of Mlc1 at the division site during cytokinesis. We also found that Myo1 plays a major role in the septin‑dependent Mlc1 localization before cytokinesis, whereas the formin Bni1 plays a major role in the actin filament–dependent Mlc1 localization during cytokinesis. Such a two‑tiered mechanism for Mlc1 localization is presumably required for the ordered assembly and robustness of cytokinesis machinery and is likely conserved across species. PMID:25631819
DOE Office of Scientific and Technical Information (OSTI.GOV)
Han, S; Ji, Y; Kim, K
Purpose: A diagnostics Multileaf Collimator (MLC) was designed for diagnostic radiography dose reduction. Monte Carlo simulation was used to evaluate efficiency of shielding material for producing leaves of Multileaf collimator. Material & Methods: The general radiography unit (Rex-650R, Listem, Korea) was modeling with Monte Carlo simulation (MCNPX, LANL, USA) and we used SRS-78 program to calculate the energy spectrum of tube voltage (80, 100, 120 kVp). The shielding materials was SKD 11 alloy tool steel that is composed of 1.6% carbon(C), 0.4% silicon (Si), 0.6% manganese (Mn), 5% chromium (Cr), 1% molybdenum (Mo), and vanadium (V). The density of itmore » was 7.89 g/m3. We simulated leafs diagnostic MLC using SKD 11 with general radiography unit. We calculated efficiency of diagnostic MLC using tally6 card of MCNPX depending on energy. Results: The diagnostic MLC consisted of 25 individual metal shielding leaves on both sides, with dimensions of 10 × 0.5 × 0.5 cm3. The leaves of MLC were controlled by motors positioned on both sides of the MLC. According to energy (tube voltage), the shielding efficiency of MLC in Monte Carlo simulation was 99% (80 kVp), 96% (100 kVp) and 93% (120 kVp). Conclusion: We certified efficiency of diagnostic MLC fabricated from SKD11 alloy tool steel. Based on the results, the diagnostic MLC was designed. We will make the diagnostic MLC for dose reduction of diagnostic radiography.« less
Automatic detection of MLC relative position errors for VMAT using the EPID-based picket fence test
NASA Astrophysics Data System (ADS)
Christophides, Damianos; Davies, Alex; Fleckney, Mark
2016-12-01
Multi-leaf collimators (MLCs) ensure the accurate delivery of treatments requiring complex beam fluences like intensity modulated radiotherapy and volumetric modulated arc therapy. The purpose of this work is to automate the detection of MLC relative position errors ⩾0.5 mm using electronic portal imaging device-based picket fence tests and compare the results to the qualitative assessment currently in use. Picket fence tests with and without intentional MLC errors were measured weekly on three Varian linacs. The picket fence images analysed covered a time period ranging between 14-20 months depending on the linac. An algorithm was developed that calculated the MLC error for each leaf-pair present in the picket fence images. The baseline error distributions of each linac were characterised for an initial period of 6 months and compared with the intentional MLC errors using statistical metrics. The distributions of median and one-sample Kolmogorov-Smirnov test p-value exhibited no overlap between baseline and intentional errors and were used retrospectively to automatically detect MLC errors in routine clinical practice. Agreement was found between the MLC errors detected by the automatic method and the fault reports during clinical use, as well as interventions for MLC repair and calibration. In conclusion the method presented provides for full automation of MLC quality assurance, based on individual linac performance characteristics. The use of the automatic method has been shown to provide early warning for MLC errors that resulted in clinical downtime.
Embryonic essential myosin light chain regulates fetal lung development in rats.
Santos, Marta; Moura, Rute S; Gonzaga, Sílvia; Nogueira-Silva, Cristina; Ohlmeier, Steffen; Correia-Pinto, Jorge
2007-09-01
Congenital diaphragmatic hernia (CDH) is currently the most life-threatening congenital anomaly the major finding of which is lung hypoplasia. Lung hypoplasia pathophysiology involves early developmental molecular insult in branching morphogenesis and a late mechanical insult by abdominal herniation in maturation and differentiation processes. Since early determinants of lung hypoplasia might appear as promising targets for prenatal therapy, proteomics analysis of normal and nitrofen-induced hypoplastic lungs was performed at 17.5 days after conception. The major differentially expressed protein was identified by mass spectrometry as myosin light chain 1a (MLC1a). Embryonic essential MLC1a and regulatory myosin light chain 2 (MLC2) were characterized throughout normal and abnormal lung development by immunohistochemistry and Western blot. Disruption of MLC1a expression was assessed in normal lung explant cultures by antisense oligodeoxynucleotides. Since early stages of normal lung development, MLC1a was expressed in vascular smooth muscle (VSM) cells of pulmonary artery, and MLC2 was present in parabronchial smooth muscle and VSM cells of pulmonary vessels. In addition, early smooth muscle differentiation delay was observed by immunohistochemistry of alpha-smooth muscle actin and transforming growth factor-beta1. Disruption of MLC1a expression during normal pulmonary development led to significant growth and branching impairment, suggesting a role in branching morphogenesis. Both MLC1a and MLC2 were absent from hypoplastic fetal lungs during pseudoglandular stage of lung development, whereas their expression partially recovered by prenatal treatment with vitamin A. Thus, a deficiency in contractile proteins MLC1a and MLC2 might have a role among the early molecular determinants of lung hypoplasia in the rat model of nitrofen-induced CDH.
Ca2+-dependent rapid Ca2+ sensitization of contraction in arterial smooth muscle.
Dimopoulos, George J; Semba, Shingo; Kitazawa, Kazuyo; Eto, Masumi; Kitazawa, Toshio
2007-01-05
Ca(2+) ion is a universal intracellular messenger that regulates numerous biological functions. In smooth muscle, Ca(2+) with calmodulin activates myosin light chain (MLC) kinase to initiate a rapid MLC phosphorylation and contraction. To test the hypothesis that regulation of MLC phosphatase is involved in the rapid development of MLC phosphorylation and contraction during Ca(2+) transient, we compared Ca(2+) signal, MLC phosphorylation, and 2 modes of inhibition of MLC phosphatase, phosphorylation of CPI-17 Thr38 and MYPT1 Thr853, during alpha(1) agonist-induced contraction with/without various inhibitors in intact rabbit femoral artery. Phenylephrine rapidly induced CPI-17 phosphorylation from a negligible amount to a peak value of 0.38+/-0.04 mol of Pi/mol within 7 seconds following stimulation, similar to the rapid time course of Ca(2+) rise and MLC phosphorylation. This rapid CPI-17 phosphorylation was dramatically inhibited by either blocking Ca(2+) release from the sarcoplasmic reticulum or by pretreatment with protein kinase C inhibitors, suggesting an involvement of Ca(2+)-dependent protein kinase C. This was followed by a slow Ca(2+)-independent and Rho-kinase/protein kinase C-dependent phosphorylation of CPI-17. In contrast, MYPT1 phosphorylation had only a slow component that increased from 0.29+/-0.09 at rest to the peak of 0.68+/-0.14 mol of Pi/mol at 1 minute, similar to the time course of contraction. Thus, there are 2 components of the Ca(2+) sensitization through inhibition of MLC phosphatase. Our results support the hypothesis that the initial rapid Ca(2+) rise induces a rapid inhibition of MLC phosphatase coincident with the Ca(2+)-induced MLC kinase activation to synergistically initiate a rapid MLC phosphorylation and contraction in arteries with abundant CPI-17 content.
Replication in the Harp File System
1981-07-01
Shrira Michael Williams iadly 1991 © Massachusetts Institute of Technology (To appear In the Proceedings of the Thirteenth ACM Symposium on Operating...S., Spector, A. Z., and Thompson, D. S. Distributed Logging for Transaction Processing. ACM Special Interest Group on Management of Data 1987 Annual ...System. USENIX Conference Proceedings , June, 1990, pp. 63-71. 15. Hagmann, R. Reimplementing the Cedar File System Using Logging and Group Commit
17 CFR 274.402 - Form ID, uniform application for access codes to file on EDGAR.
Code of Federal Regulations, 2010 CFR
2010-04-01
... for access codes to file on EDGAR. 274.402 Section 274.402 Commodity and Securities Exchanges... Forms for Electronic Filing § 274.402 Form ID, uniform application for access codes to file on EDGAR..., filing agent or training agent to log on to the EDGAR system, submit filings, and change its CCC. (d...
17 CFR 274.402 - Form ID, uniform application for access codes to file on EDGAR.
Code of Federal Regulations, 2011 CFR
2011-04-01
... for access codes to file on EDGAR. 274.402 Section 274.402 Commodity and Securities Exchanges... Forms for Electronic Filing § 274.402 Form ID, uniform application for access codes to file on EDGAR..., filing agent or training agent to log on to the EDGAR system, submit filings, and change its CCC. (d...
NASA Astrophysics Data System (ADS)
Shiinoki, Takehiro; Hanazawa, Hideki; Yuasa, Yuki; Fujimoto, Koya; Uehara, Takuya; Shibuya, Keiko
2017-02-01
A combined system comprising the TrueBeam linear accelerator and a new real-time tumour-tracking radiotherapy system, SyncTraX, was installed at our institution. The objectives of this study are to develop a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine electronic portal image device (EPID) images and a log file and to verify this treatment in clinical cases. Respiratory-gated radiotherapy was performed using TrueBeam and the SyncTraX system. Cine EPID images and a log file were acquired for a phantom and three patients during the course of the treatment. Digitally reconstructed radiographs (DRRs) were created for each treatment beam using a planning CT set. The cine EPID images, log file, and DRRs were analysed using a developed software. For the phantom case, the accuracy of the proposed method was evaluated to verify the respiratory-gated radiotherapy. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker used as an internal surrogate were calculated to evaluate the gating accuracy and set-up uncertainty in the superior-inferior (SI), anterior-posterior (AP), and left-right (LR) directions. The proposed method achieved high accuracy for the phantom verification. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker were ⩽3 mm and ±3 mm in the SI, AP, and LR directions. We proposed a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine EPID images and a log file and showed that this treatment is performed with high accuracy in clinical cases. This work was partly presented at the 58th Annual meeting of American Association of Physicists in Medicine.
Shiinoki, Takehiro; Hanazawa, Hideki; Yuasa, Yuki; Fujimoto, Koya; Uehara, Takuya; Shibuya, Keiko
2017-02-21
A combined system comprising the TrueBeam linear accelerator and a new real-time tumour-tracking radiotherapy system, SyncTraX, was installed at our institution. The objectives of this study are to develop a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine electronic portal image device (EPID) images and a log file and to verify this treatment in clinical cases. Respiratory-gated radiotherapy was performed using TrueBeam and the SyncTraX system. Cine EPID images and a log file were acquired for a phantom and three patients during the course of the treatment. Digitally reconstructed radiographs (DRRs) were created for each treatment beam using a planning CT set. The cine EPID images, log file, and DRRs were analysed using a developed software. For the phantom case, the accuracy of the proposed method was evaluated to verify the respiratory-gated radiotherapy. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker used as an internal surrogate were calculated to evaluate the gating accuracy and set-up uncertainty in the superior-inferior (SI), anterior-posterior (AP), and left-right (LR) directions. The proposed method achieved high accuracy for the phantom verification. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker were ⩽3 mm and ±3 mm in the SI, AP, and LR directions. We proposed a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine EPID images and a log file and showed that this treatment is performed with high accuracy in clinical cases.
Workload Characterization and Performance Implications of Large-Scale Blog Servers
DOE Office of Scientific and Technical Information (OSTI.GOV)
Jeon, Myeongjae; Kim, Youngjae; Hwang, Jeaho
With the ever-increasing popularity of social network services (SNSs), an understanding of the characteristics of these services and their effects on the behavior of their host servers is critical. However, there has been a lack of research on the workload characterization of servers running SNS applications such as blog services. To fill this void, we empirically characterized real-world web server logs collected from one of the largest South Korean blog hosting sites for 12 consecutive days. The logs consist of more than 96 million HTTP requests and 4.7 TB of network traffic. Our analysis reveals the followings: (i) The transfermore » size of non-multimedia files and blog articles can be modeled using a truncated Pareto distribution and a log-normal distribution, respectively; (ii) User access for blog articles does not show temporal locality, but is strongly biased towards those posted with image or audio files. We additionally discuss the potential performance improvement through clustering of small files on a blog page into contiguous disk blocks, which benefits from the observed file access patterns. Trace-driven simulations show that, on average, the suggested approach achieves 60.6% better system throughput and reduces the processing time for file access by 30.8% compared to the best performance of the Ext4 file system.« less
Ryder, Robert T.; Swezey, Christopher S.; Crangle, Robert D.; Trippi, Michael H.; Ruppert, Leslie F.; Ryder, Robert T.
2014-01-01
This chapter is a re-release of U.S. Geological Survey Scientific Investigations Map 2985, of the same title, by Ryder and others (2008). For this chapter, two appendixes have been added that do not appear with the original version. Appendix A provides Log ASCII Standard (LAS) files for each drill hole along cross-section E–E'; they are text files which encode gamma-ray, neutron, density, and other logs that can be used by most well-logging software. Appendix B provides graphic well-log traces from each drill hole.
Liu, Shi; Wu, Yu; Wooten, H Omar; Green, Olga; Archer, Brent; Li, Harold; Yang, Deshan
2016-03-08
A software tool is developed, given a new treatment plan, to predict treatment delivery time for radiation therapy (RT) treatments of patients on ViewRay magnetic resonance image-guided radiation therapy (MR-IGRT) delivery system. This tool is necessary for managing patient treatment scheduling in our clinic. The predicted treatment delivery time and the assessment of plan complexities could also be useful to aid treatment planning. A patient's total treatment delivery time, not including time required for localization, is modeled as the sum of four components: 1) the treatment initialization time; 2) the total beam-on time; 3) the gantry rotation time; and 4) the multileaf collimator (MLC) motion time. Each of the four components is predicted separately. The total beam-on time can be calculated using both the planned beam-on time and the decay-corrected dose rate. To predict the remain-ing components, we retrospectively analyzed the patient treatment delivery record files. The initialization time is demonstrated to be random since it depends on the final gantry angle of the previous treatment. Based on modeling the relationships between the gantry rotation angles and the corresponding rotation time, linear regression is applied to predict the gantry rotation time. The MLC motion time is calculated using the leaves delay modeling method and the leaf motion speed. A quantitative analysis was performed to understand the correlation between the total treatment time and the plan complexity. The proposed algorithm is able to predict the ViewRay treatment delivery time with the average prediction error 0.22min or 1.82%, and the maximal prediction error 0.89 min or 7.88%. The analysis has shown the correlation between the plan modulation (PM) factor and the total treatment delivery time, as well as the treatment delivery duty cycle. A possibility has been identified to significantly reduce MLC motion time by optimizing the positions of closed MLC pairs. The accuracy of the proposed prediction algorithm is sufficient to support patient treatment appointment scheduling. This developed software tool is currently applied in use on a daily basis in our clinic, and could also be used as an important indicator for treatment plan complexity.
SU-E-T-629: Prediction of the ViewRay Radiotherapy Treatment Time for Clinical Logistics
DOE Office of Scientific and Technical Information (OSTI.GOV)
Liu, S; Wooten, H; Wu, Y
Purpose: An algorithm is developed in our clinic, given a new treatment plan, to predict treatment delivery time for radiation therapy (RT) treatments of patients on ViewRay magnetic resonance-image guided radiation therapy (MR-IGRT) delivery system. This algorithm is necessary for managing patient treatment appointments, and is useful as an indicator to assess the treatment plan complexity. Methods: A patient’s total treatment delivery time, not including time required for localization, may be described as the sum of four components: (1) the treatment initialization time; (2) the total beam-on time; (3) the gantry rotation time; and (4) the multileaf collimator (MLC) motionmore » time. Each of the four components is predicted separately. The total beam-on time can be calculated using both the planned beam-on time and the decay-corrected delivery dose rate. To predict the remaining components, we quantitatively analyze the patient treatment delivery record files. The initialization time is demonstrated to be random since it depends on the final gantry angle and MLC leaf positions of the previous treatment. Based on modeling the relationships between the gantry rotation angles and the corresponding rotation time, and between the furthest MLC leaf moving distance and the corresponding MLC motion time, the total delivery time is predicted using linear regression. Results: The proposed algorithm has demonstrated the feasibility of predicting the ViewRay treatment delivery time for any treatment plan of any patient. The average prediction error is 0.89 minutes or 5.34%, and the maximal prediction error is 2.09 minutes or 13.87%. Conclusion: We have developed a treatment delivery time prediction algorithm based on the analysis of previous patients’ treatment delivery records. The accuracy of our prediction is sufficient for guiding and arranging patient treatment appointments on a daily basis. The predicted delivery time could also be used as an indicator to assess the treatment plan complexity. This work was supported by a research grant from Viewray Inc.« less
INSPIRE and SPIRES Log File Analysis
DOE Office of Scientific and Technical Information (OSTI.GOV)
Adams, Cole; /Wheaton Coll. /SLAC
2012-08-31
SPIRES, an aging high-energy physics publication data base, is in the process of being replaced by INSPIRE. In order to ease the transition from SPIRES to INSPIRE it is important to understand user behavior and the drivers for adoption. The goal of this project was to address some questions in regards to the presumed two-thirds of the users still using SPIRES. These questions are answered through analysis of the log files from both websites. A series of scripts were developed to collect and interpret the data contained in the log files. The common search patterns and usage comparisons are mademore » between INSPIRE and SPIRES, and a method for detecting user frustration is presented. The analysis reveals a more even split than originally thought as well as the expected trend of user transition to INSPIRE.« less
Federal Register 2010, 2011, 2012, 2013, 2014
2013-07-05
.... Select the eFiling link to log on and submit the intervention or protests. Persons unable to file... to protest should file with the Federal Energy Regulatory Commission, 888 First Street NE... facilitate electronic service, persons with Internet access who will eFile a document and/or be listed as a...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-06-07
.... Select the eFiling link to log on and submit the intervention or protests. Persons unable to file... to protest should file with the Federal Energy Regulatory Commission, 888 First Street NE... facilitate electronic service, persons with Internet access who will eFile a document and/or be listed as a...
SU-E-T-444: Gravity Effect On Maximum Leaf Speed in Dynamic IMRT Treatments
DOE Office of Scientific and Technical Information (OSTI.GOV)
Olasolo, J; Pellejero, S; Gracia, M
Purpose: A leaf sequencing algorithm has been recently developed in our department. Our purpose is to utilize this algorithm to reduce treatment time by studying the feasibility of using several maximum leaf speeds depending on gantry angle and leaf thickness (0.5 or 1 cm at isocenter). To do so, the gravity effect on MLC performance has been examined by means of analysing the dynalog files. Methods: Leaf position errors has been ascertained according to gantry angle and leaf speed in MLC Millenium120 (Varian). In order to do this, the following test has been designed: all leaves move in synchrony, withmore » same speed and 1 cm gap between opposite leaves. This test is implemented for 18 different speeds: 0.25-0.5-0.75-1-1.25-1.5-1.75-2-2.1-2.2-2.3-2.4-2.5-2.6-2.7-2.8-2.9-3.0 cm/s and 8 gantry angles: 0-45-90-135-180-225-270-315. Collimator angle is 2 degrees in all cases since it is the most usual one in IMRT treatments in our department. Dynamic tolerance is 2 mm. Dynalogs files of 10 repetitions of the test are analysed with a Mathlab in-house developed software and RMS error and 95th percentiles are calculated. Varian recommends 2.5 cm/s as the maximum leaf speed for its segmentation algorithm. In our case, we accept this speed in the most restrictive situation: gantry angle 270 and 1 cm leaf thickness. Maximum speeds for the rest of the cases are calculated by keeping the difference between 95th percentile and dynamic tolerance. In this way, beam hold-off probability does not increase. Results: Maximum speeds every 45 degrees of gantry rotation have been calculated for both leaf thickness. These results are 2.9-2.9-2.9-2.9-2.7-2.6-2.6-2.7 cm/s for 0.5 cm leaf thickness and 2.7-2.7-2.7-2.7-2.6-2.5-2.5-2.6 cm/s for 1 cm leaf thickness. Conclusion: Gravity effect on MLC positioning has been studied. Maximum leaf speed according to leaf thickness and gantry angle have been calculated which reduces treatment time.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Lin Yu; Wang Fuji; Tao Yan
2000-07-01
This paper introduced a new idea of transporting mine tailings-logs in mine tailings-slurry pipeline and a new technology of mine cemented filing of tailings-logs with tailings-slurry. The hydraulic principles, the compaction of tailings-logs and the mechanic function of fillbody of tailings-logs cemented by tailings-slurry have been discussed.
Study of the IMRT interplay effect using a 4DCT Monte Carlo dose calculation.
Jensen, Michael D; Abdellatif, Ady; Chen, Jeff; Wong, Eugene
2012-04-21
Respiratory motion may lead to dose errors when treating thoracic and abdominal tumours with radiotherapy. The interplay between complex multileaf collimator patterns and patient respiratory motion could result in unintuitive dose changes. We have developed a treatment reconstruction simulation computer code that accounts for interplay effects by combining multileaf collimator controller log files, respiratory trace log files, 4DCT images and a Monte Carlo dose calculator. Two three-dimensional (3D) IMRT step-and-shoot plans, a concave target and integrated boost were delivered to a 1D rigid motion phantom. Three sets of experiments were performed with 100%, 50% and 25% duty cycle gating. The log files were collected, and five simulation types were performed on each data set: continuous isocentre shift, discrete isocentre shift, 4DCT, 4DCT delivery average and 4DCT plan average. Analysis was performed using 3D gamma analysis with passing criteria of 2%, 2 mm. The simulation framework was able to demonstrate that a single fraction of the integrated boost plan was more sensitive to interplay effects than the concave target. Gating was shown to reduce the interplay effects. We have developed a 4DCT Monte Carlo simulation method that accounts for IMRT interplay effects with respiratory motion by utilizing delivery log files.
User's manual for SEDCALC, a computer program for computation of suspended-sediment discharge
Koltun, G.F.; Gray, John R.; McElhone, T.J.
1994-01-01
Sediment-Record Calculations (SEDCALC), a menu-driven set of interactive computer programs, was developed to facilitate computation of suspended-sediment records. The programs comprising SEDCALC were developed independently in several District offices of the U.S. Geological Survey (USGS) to minimize the intensive labor associated with various aspects of sediment-record computations. SEDCALC operates on suspended-sediment-concentration data stored in American Standard Code for Information Interchange (ASCII) files in a predefined card-image format. Program options within SEDCALC can be used to assist in creating and editing the card-image files, as well as to reformat card-image files to and from formats used by the USGS Water-Quality System. SEDCALC provides options for creating card-image files containing time series of equal-interval suspended-sediment concentrations from 1. digitized suspended-sediment-concentration traces, 2. linear interpolation between log-transformed instantaneous suspended-sediment-concentration data stored at unequal time intervals, and 3. nonlinear interpolation between log-transformed instantaneous suspended-sediment-concentration data stored at unequal time intervals. Suspended-sediment discharge can be computed from the streamflow and suspended-sediment-concentration data or by application of transport relations derived by regressing log-transformed instantaneous streamflows on log-transformed instantaneous suspended-sediment concentrations or discharges. The computed suspended-sediment discharge data are stored in card-image files that can be either directly imported to the USGS Automated Data Processing System or used to generate plots by means of other SEDCALC options.
Li, Yinghui; Chen, Lixin; Zhu, Jinhan; Wang, Bin; Liu, Xiaowei
2017-07-01
A quantitative method based on the electronic portal imaging system (EPID) and film was developed for MLC position and speed testing; this method was used for three MLC types (Millennium, MLCi, and Agility MLC). To determine the leaf position, a picket fence designed by the dynamic (DMLC) model was used. The full-width half-maximum (FWHM) values of each gap measured by EPID and EBT3 were converted to the gap width using the FWHM versus nominal gap width relationship. The algorithm developed for the picket fence analysis was able to quantify the gap width, the distance between gaps, and each individual leaf position. To determine the leaf speed, a 0.5 × 20 cm 2 MLC-defined sliding gap was applied across a 14 × 20 cm 2 symmetry field. The linacs ran at a fixed-dose rate. The use of different monitor units (MUs) for this test led to different leaf speeds. The effect of leaf transmission was considered in a speed accuracy analysis. The difference between the EPID and film results for the MLC position is less than 0.1 mm. For the three MLC types, twice the standard deviation (2 SD) is provided; 0.2, 0.4, and 0.4 mm for gap widths of three MLC types, and 0.1, 0.2, and 0.2 mm for distances between gaps. The individual leaf positions deviate from the preset positions within 0.1 mm. The variations in the speed profiles for the EPID and EBT3 results are consistent, but the EPID results are slightly better than the film results. Different speeds were measured for each MLC type. For all three MLC types, speed errors increase with increasing speed. The analysis speeds deviate from the preset speeds within approximately 0.01 cm s -1 . This quantitative analysis of MLC position and speed provides an intuitive evaluation for MLC quality assurance (QA). © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Hossain, S; Keeling, V; Ahmad, S
Purpose: To determine the effects of multileaf collimator (MLC) leaf width on normal-brain-tissue doses and dose conformity of SRS RapidArc treatment plans for brain tumors. Methods: Ten patients with 24 intracranial tumors (seven with 1–2 and three with 4–6 lesions) were planned using RapidArc for both Varian Millennium 120 MLC (5 mm leaf width) and high definition (HD) MLC (2.5 mm leaf width). Between 2 and 8 arcs were used with two full coplanar arcs and the rest non-coplanar half arcs. 6 MV beams were used and plans were optimized with a high priority to the Normal Tissue Objective (tomore » achieve dose conformity and sharp dose fall-off) and normal brain tissue. Calculation was done using AAA on a 1 mm grid size. The prescription dose ranged from 14–22 Gy. Plans were normalized such that 99% of the target received the prescription dose. Identical beam geometries, optimizations, calculations, and normalizations were used for both plans. Paddick Conformity Index (PCI), V4, V8 and V12 Gy for normal brain tissue and Integral Dose were used for analysis. Results: In all cases, HD MLC plans performed better in sparing normal brain tissue, achieving a higher PCI with a lower Integral Dose. The average PCI for all 24 targets was 0.75±0.23 and 0.70±0.23 (p ≤0.0015) for HD MLC and Millennium MLC plans, respectively. The average ratio of normal brain doses for Millennium MLC to HD MLC plans was 1.30±0.16, 1.27±0.15, and 1.31±0.18 for the V4, V8, and V12, respectively. The differences in normal brain dose for all criteria were statistically significant with p-value < 0.02. On average Millennium MLC plans had a 16% higher integral dose than HD MLC plans. Conclusion: Significantly better dose conformity with reduced volume of normal brain tissue and integral dose was achieved with HD MLC plans compared to Millennium MLC plans.« less
SU-G-BRA-14: Dose in a Rigidly Moving Phantom with Jaw and MLC Compensation
DOE Office of Scientific and Technical Information (OSTI.GOV)
Chao, E; Lucas, D
Purpose: To validate dose calculation for a rigidly moving object with jaw motion and MLC shifts to compensate for the motion in a TomoTherapy™ treatment delivery. Methods: An off-line version of the TomoTherapy dose calculator was extended to perform dose calculations for rigidly moving objects. A variety of motion traces were added to treatment delivery plans, along with corresponding jaw compensation and MLC shift compensation profiles. Jaw compensation profiles were calculated by shifting the jaws such that the center of the treatment beam moved by an amount equal to the motion in the longitudinal direction. Similarly, MLC compensation profiles weremore » calculated by shifting the MLC leaves by an amount that most closely matched the motion in the transverse direction. The same jaw and MLC compensation profiles were used during simulated treatment deliveries on a TomoTherapy system, and film measurements were obtained in a rigidly moving phantom. Results: The off-line TomoTherapy dose calculator accurately predicted dose profiles for a rigidly moving phantom along with jaw motion and MLC shifts to compensate for the motion. Calculations matched film measurements to within 2%/1 mm. Jaw and MLC compensation substantially reduced the discrepancy between the delivered dose distribution and the calculated dose with no motion. For axial motion, the compensated dose matched the no-motion dose within 2%/1mm. For transverse motion, the dose matched within 2%/3mm (approximately half the width of an MLC leaf). Conclusion: The off-line TomoTherapy dose calculator accurately computes dose delivered to a rigidly moving object, and accurately models the impact of moving the jaws and shifting the MLC leaf patterns to compensate for the motion. Jaw tracking and MLC leaf shifting can effectively compensate for the dosimetric impact of motion during a TomoTherapy treatment delivery.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Ding, C; Timmerman, R; Jiang, S
Purpose: To evaluate the dosimetric impact on treatment planning for partial breast stereotactic irradiation using Cyberknife with MLC versus Iris Cone. Methods: Ten patients whom underwent lumpectomy for DCIS or stage I invasive non-lobular epithelial breast cancer were included in this study. All patients were previously treated on the Cyberknife using Iris cone with the prescription dose of 37.5Gy in 5 fractions covering at least 95% of PTV on our phase I SBRT 5 fraction partial breast irradiation trial. Retrospectively, treatment planning was performed and compared using the new Cyberknife M6 MLC system for each patient. Using the same contoursmore » and critical organ constraints for both MLC and Iris cone plans, the dose on target and critical organs were analyzed accordingly. Results: Dose to critical organs such as ipsilateral lung, contralateral lung, heart, skin, ipsilateral breast, and rib were analyzed, as well as conformity index and high dose spillage of the target area. In 9 of 10 patients, the MLC plans had less total ipsilateral breast volume encompassing the 50% prescription isodose (mean:22.3±8.2% MLC vs. 31.6±8.0 Iris, p=0.00014) .The MLC plans mean estimated treatment delivery time was significantly less than the Iris plans (51±3.9min vs. 56.2±9min, p=0.03) Both MLC and Iris cone plans were able to meet all dose constraints and there was no statistical difference between those dose constraints. Conclusion: Both MLC and Iris Cone can deliver conformal dose to a partial breast target and satisfy the dose constraints of critical organs. The new Cyberknife with MLC can deliver a more conformal dose in the lower dose region and spare more ipsilateral breast tissue to the 50% prescription isodose. The treatment time for partial breast SBRT plans was also reduced using MLC. Project receives research support from Accuray Inc.« less
Saito, Masahide; Sano, Naoki; Shibata, Yuki; Kuriyama, Kengo; Komiyama, Takafumi; Marino, Kan; Aoki, Shinichi; Ashizawa, Kazunari; Yoshizawa, Kazuya; Onishi, Hiroshi
2018-05-01
The purpose of this study was to compare the MLC error sensitivity of various measurement devices for VMAT pre-treatment quality assurance (QA). This study used four QA devices (Scandidos Delta4, PTW 2D-array, iRT systems IQM, and PTW Farmer chamber). Nine retrospective VMAT plans were used and nine MLC error plans were generated for all nine original VMAT plans. The IQM and Farmer chamber were evaluated using the cumulative signal difference between the baseline and error-induced measurements. In addition, to investigate the sensitivity of the Delta4 device and the 2D-array, global gamma analysis (1%/1, 2%/2, and 3%/3 mm), dose difference (1%, 2%, and 3%) were used between the baseline and error-induced measurements. Some deviations of the MLC error sensitivity for the evaluation metrics and MLC error ranges were observed. For the two ionization devices, the sensitivity of the IQM was significantly better than that of the Farmer chamber (P < 0.01) while both devices had good linearly correlation between the cumulative signal difference and the magnitude of MLC errors. The pass rates decreased as the magnitude of the MLC error increased for both Delta4 and 2D-array. However, the small MLC error for small aperture sizes, such as for lung SBRT, could not be detected using the loosest gamma criteria (3%/3 mm). Our results indicate that DD could be more useful than gamma analysis for daily MLC QA, and that a large-area ionization chamber has a greater advantage for detecting systematic MLC error because of the large sensitive volume, while the other devices could not detect this error for some cases with a small range of MLC error. © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
SU-E-T-646: Quality Assurance of Truebeam Multi-Leaf Collimator Using a MLC QA Phantom
DOE Office of Scientific and Technical Information (OSTI.GOV)
Zhang, J; Lu, J; Hong, D
2015-06-15
Purpose: To perform a routine quality assurance procedure for Truebeam multi-leaf collimator (MLC) using MLC QA phantom, verify the stability and reliability of MLC during the treatment. Methods: MLC QA phantom is a specialized phantom for MLC quality assurance (QA), and contains five radio-opaque spheres that are embedded in an “L” shape. The phantom was placed isocentrically on the Truebeam treatment couch for the tests. A quality assurance plan was setted up in the Eclipse v10.0, the fields that need to be delivered in order to acquire the necessary images, the MLC shapes can then be obtained by the images.more » The images acquired by the electronic portal imaging device (EPID), and imported into the PIPSpro software for the analysis. The tests were delivered twelve weeks (once a week) to verify consistency of the delivery, and the images are acquired in the same manner each time. Results: For the Leaf position test, the average position error was 0.23mm±0.02mm (range: 0.18mm∼0.25mm). The Leaf width was measured at the isocenter, the average error was 0.06mm±0.02mm (range: 0.02mm∼0.08mm) for the Leaf width test. Multi-Port test showed the dynamic leaf shift error, the average error was 0.28mm±0.03mm (range: 0.2mm∼0.35mm). For the leaf transmission test, the average inter-leaf leakage value was 1.0%±0.17% (range: 0.8%∼1.3%) and the average inter-bank leakage value was 32.6%±2.1% (range: 30.2%∼36.1%). Conclusion: By the test of 12 weeks, the MLC system of the Truebeam is running in a good condition and the MLC system can be steadily and reliably carried out during the treatment. The MLC QA phantom is a useful test tool for the MLC QA.« less
TH-AB-BRA-01: A Novel Doubly-Focused Multileaf Collimator Design for MR-Guided Radiation Therapy
DOE Office of Scientific and Technical Information (OSTI.GOV)
Li, H; Mutic, S; Green, O
2016-06-15
Purpose: To describe the physical and dosimetric properties of a novel double-stack multileaf collimator (MLC). Methods: One of the compromises made in the MLC design has been to employ linear-motion singly-divergent shapes. Because the MLC leading edge moves linearly, it is rounded to provide a consistent, albeit compromised penumbra. The MLC employed in the new linac-based MR-IGRT unit is designed to be doubly focused in that each leaf moves in an arc centered at the source, and the sides of the leaves are machined such that they lie parallel to a line between the leaf edge and the source. Themore » curvature of the MLC keeps motors and encoders in lower magnetic field. However, high spatial-resolution leaves are difficult to manufacture to sufficiently tight tolerances and difficult to move due to restricted space on the gantry. Wider leaves alleviate this problem with less moving parts but the coarse resolution disallows treating very small lesions. This compromise has been overcome by splitting the MLC leaf bank into two sets, stacked one upon the other and offset half of a leaf width. The dosimetry has been simulated using Monte-Carlo and a 6 MV linac in a 0.35 T magnetic field. Results: The combined MLC leaf set has a spatial resolution of effectively half of the leaf width, 4mm here. The dosimetry resolution and conformality are consistent with 4mm wide MLC assisted by inverse fluence modulation. Also, because each leaf junction is backed up by the stacked leaf that lies over the junction, the problem of tongue-and-groove dosimetry has been greatly reduced. The novel MLC design allows the use of more powerful leaf motors than would be otherwise possible if a single MLC bank is employed. Conclusions: The stacked MLC will provide highly conformal dose distributions suitable for stereotactic radiation therapy of small lesions. The research was funded by ViewRay, Inc.« less
Childress, Catherine; Feuerbacher, Leigh A.; Phillips, Linda; Burgum, Alex
2013-01-01
Aggregatibacter actinomycetemcomitans, a periodontal pathogen, synthesizes leukotoxin (LtxA), a protein that helps the bacterium evade the host immune response. Transcription of the ltxA operon is induced during anaerobic growth. The cyclic AMP (cAMP) receptor protein (CRP) indirectly increases ltxA expression, but the intermediary regulator is unknown. Integration host factor (IHF) binds to and represses the leukotoxin promoter, but neither CRP nor IHF is responsible for the anaerobic induction of ltxA RNA synthesis. Thus, we have undertaken studies to identify other regulators of leukotoxin transcription and to demonstrate how these proteins work together to modulate leukotoxin synthesis. First, analyses of ltxA RNA expression from defined leukotoxin promoter mutations in the chromosome identify positions −69 to −35 as the key control region and indicate that an activator protein modulates leukotoxin transcription. We show that Mlc, which is a repressor in Escherichia coli, functions as a direct transcriptional activator in A. actinomycetemcomitans; an mlc deletion mutant reduces leukotoxin RNA synthesis, and recombinant Mlc protein binds specifically at the −68 to −40 region of the leukotoxin promoter. Furthermore, we show that CRP activates ltxA expression indirectly by increasing the levels of Mlc. Analyses of Δmlc, Δihf, and Δihf Δmlc strains demonstrate that Mlc can increase RNA polymerase (RNAP) activity directly and that IHF represses ltxA RNA synthesis mainly by blocking Mlc binding. Finally, a Δihf Δmlc mutant still induces ltxA during anaerobic growth, indicating that there are additional factors involved in leukotoxin transcriptional regulation. A model for the coordinated regulation of leukotoxin transcription is presented. PMID:23475968
SU-E-T-501: Initial Orthovoltage Beam Profile Analysis of a Small Brass MLC
DOE Office of Scientific and Technical Information (OSTI.GOV)
Loughery, B; Snyder, M
2015-06-15
Purpose To create brass leaves for an orthovoltage MLC and take initial beam profile measurements. Methods The low-energy MLC was designed in previous work. Brass was chosen for its self-lubrication and low cost. Stock brass rectangles (30cm × 1.0cm × 0.5cm) were ordered with pre-cut gear rack along the topmost long edges. Leaf designs were translated into G-code, then cut with a Tormach CNC-1100 mill. Intense bowing was observed in the beam direction, which required straightening via an in-house jig. Straightened leaves were placed into MLC assembly and mounted to a 320 kVp orthovoltage tube. EDR2 film was irradiated inmore » four situations: MLC open so one edge was isocentric, and MLC open more than isocentric, completely closed MLC, and an open field shot with the MLC removed. The first two scans tested penumbra for our rectangular edges due to unfocused design. The final two scans tested transmission and interleaf leakage. All four experiments were set to 120 kVp and 10 mA for two minutes. Results Transmission and interleaf leakage were found to be zero. Interleaf leakage is faintly visible on film, but undetected by our film scanner despite high spatial resolution. Penumbra at isocenter was found to be 0.72mm, which matched the penumbras of true field edges. Penumbra off-isocenter was 1.1mm. Mechanically, leaves are moving smoothly once straightened. Conclusion Beam profiles through our brass MLC are acceptable. Leaves attenuate and move as designed. Looking forward, we intend to animate our MLC to deliver more complicated treatment plans.« less
Al-Shboul, Othman A.; Al-Dwairi, Ahmed N.; Alqudah, Mohammad A.; Mustafa, Ayman G.
2018-01-01
Evidence of sex-related differences in gastrointestinal (GI) functions has been reported in the literature. In addition, various GI disorders have disproportionate prevalence between the sexes. An essential step in the initiation of smooth muscle contraction is the phosphorylation of the 20-kDa regulatory myosin light chain (MLC20) by the Ca2+/calmodulin-dependent myosin light chain kinase (MLCK). However, whether male stomach smooth muscle inherits different contractile signaling mechanisms for the regulation of MLC20 phosphorylation from that in females has not been established. The present study was designed to investigate sex-associated differences in the regulation of MLC20 phosphorylation and thus muscle contraction in gastric smooth muscle cells (GSMCs). Experiments were performed on GSMCs freshly isolated from male and female rats. Contraction of the GSMCs in response to acetylcholine (ACh), a muscarinic agonist, was measured via scanning micrometry in the presence or absence of the MLCK inhibitor, ML-7. Additionally, the protein levels of MLC20, MLCK and phosphorylated MLC20 were measured by ELISA. The protein levels of MLC20 and MLCK were indifferent between the sexes. ACh induced greater contraction (P<0.05) as well as greater MLC20 phosphorylation (P<0.05) in male GSMCs compared with female. Pretreatment of GSMCs with ML-7 significantly reduced the ACh-induced contraction (P<0.05) and MLC20 phosphorylation (P<0.05) in the male and female cells, and notably, abolished the contractile differences between the sexes. In conclusion, MLC20 phosphorylation and thus muscle contraction may be activated to a greater extent in male rat stomach compared with that in females. PMID:29599980
DOE Office of Scientific and Technical Information (OSTI.GOV)
Bai, Sen; Li, Guangjun; Wang, Maojie
The purpose of this study was to investigate the effect of multileaf collimator (MLC) leaf position, collimator rotation angle, and accelerator gantry rotation angle errors on intensity-modulated radiotherapy plans for nasopharyngeal carcinoma. To compare dosimetric differences between the simulating plans and the clinical plans with evaluation parameters, 6 patients with nasopharyngeal carcinoma were selected for simulation of systematic and random MLC leaf position errors, collimator rotation angle errors, and accelerator gantry rotation angle errors. There was a high sensitivity to dose distribution for systematic MLC leaf position errors in response to field size. When the systematic MLC position errors weremore » 0.5, 1, and 2 mm, respectively, the maximum values of the mean dose deviation, observed in parotid glands, were 4.63%, 8.69%, and 18.32%, respectively. The dosimetric effect was comparatively small for systematic MLC shift errors. For random MLC errors up to 2 mm and collimator and gantry rotation angle errors up to 0.5°, the dosimetric effect was negligible. We suggest that quality control be regularly conducted for MLC leaves, so as to ensure that systematic MLC leaf position errors are within 0.5 mm. Because the dosimetric effect of 0.5° collimator and gantry rotation angle errors is negligible, it can be concluded that setting a proper threshold for allowed errors of collimator and gantry rotation angle may increase treatment efficacy and reduce treatment time.« less
Andruchov, Oleg; Galler, Stefan
2008-03-01
This study contributes to understand the physiological role of slow myosin light chain isoforms in fast-twitch type IIA fibres of skeletal muscle. These isoforms are often attached to the myosin necks of rat type IIA fibres, whereby the slow alkali myosin light chain isoform MLC1s is much more frequent and abundant than the slow regulatory myosin light chain isoform MLC2s. In the present study, single-skinned rat type IIA fibres were maximally Ca(2+) activated and subjected to stepwise stretches for causing a perturbation of myosin head pulling cycles. From the time course of the resulting force transients, myosin head kinetics was deduced. Fibres containing MLC1s exhibited slower kinetics independently of the presence or absence of MLC2s. At the maximal MLC1s concentration of about 75%, the slowing was about 40%. The slowing effect of MLC1s is possibly due to differences in the myosin heavy chain binding sites of the fast and slow alkali MLC isoforms, which changes the rigidity of the myosin neck. Compared with the impact of myosin heavy chain isoforms in various fast-twitch fibre types, the influence of MLC1s on myosin head kinetics of type IIA fibres is much smaller. In conclusion, the physiological role of fast and slow MLC isoforms in type IIA fibres is a fine-tuning of the myosin head kinetics.
17 CFR 239.63 - Form ID, uniform application for access codes to file on EDGAR.
Code of Federal Regulations, 2011 CFR
2011-04-01
... for access codes to file on EDGAR. 239.63 Section 239.63 Commodity and Securities Exchanges SECURITIES... Statements § 239.63 Form ID, uniform application for access codes to file on EDGAR. Form ID must be filed by... log on to the EDGAR system, submit filings, and change its CCC. (d) Password Modification...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-09-06
... the eFiling link to log on and submit the intervention or protests. Persons unable to file... protest should file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC... electronic service, persons with Internet access who will eFile a document and/or be listed as a contact for...
17 CFR 239.63 - Form ID, uniform application for access codes to file on EDGAR.
Code of Federal Regulations, 2010 CFR
2010-04-01
... for access codes to file on EDGAR. 239.63 Section 239.63 Commodity and Securities Exchanges SECURITIES... Statements § 239.63 Form ID, uniform application for access codes to file on EDGAR. Form ID must be filed by... log on to the EDGAR system, submit filings, and change its CCC. (d) Password Modification...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-05-16
.... Select the eFiling link to log on and submit the intervention or protests. Persons unable to file... intervene or to protest should file with the Federal Energy Regulatory Commission, 888 First Street NE... facilitate electronic service, persons with Internet access who will eFile a document and/or be listed as a...
77 FR 55817 - Delek Crude Logistics, LLC; Notice of Petition for Waiver
Federal Register 2010, 2011, 2012, 2013, 2014
2012-09-11
... using the eRegistration link. Select the eFiling link to log on and submit the intervention or protests... number. eFiling is encouraged. More detailed information relating to filing requirements, interventions...'') grant a temporary waiver of the filing and reporting requirements of sections 6 and 201 of the...
Ryder, Robert T.; Crangle, Robert D.; Trippi, Michael H.; Swezey, Christopher S.; Lentz, Erika E.; Rowan, Elisabeth L.; Hope, Rebecca S.; Ruppert, Leslie F.; Ryder, Robert T.
2014-01-01
This chapter is a re-release of U.S. Geological Survey Scientific Investigations Map 3067, of the same title, by Ryder and others (2009). For this chapter, two appendixes have been added that do not appear with the original version. Appendix A provides Log ASCII Standard (LAS) files for each drill hole along cross-section D-D'; they are text files which encode gamma-ray, neutron, density, and other logs that can be used by most well-logging software. Appendix B provides graphic well-log traces and lithologic descriptions with formation tops from each drill hole.
Parallel checksumming of data chunks of a shared data object using a log-structured file system
Bent, John M.; Faibish, Sorin; Grider, Gary
2016-09-06
Checksum values are generated and used to verify the data integrity. A client executing in a parallel computing system stores a data chunk to a shared data object on a storage node in the parallel computing system. The client determines a checksum value for the data chunk; and provides the checksum value with the data chunk to the storage node that stores the shared object. The data chunk can be stored on the storage node with the corresponding checksum value as part of the shared object. The storage node may be part of a Parallel Log-Structured File System (PLFS), and the client may comprise, for example, a Log-Structured File System client on a compute node or burst buffer. The checksum value can be evaluated when the data chunk is read from the storage node to verify the integrity of the data that is read.
15 CFR 762.3 - Records exempt from recordkeeping requirements.
Code of Federal Regulations, 2010 CFR
2010-01-01
...; (2) Special export file list; (3) Vessel log from freight forwarder; (4) Inspection certificate; (5... form; (12) Financial hold form; (13) Export parts shipping problem form; (14) Draft number log; (15) Expense invoice mailing log; (16) Financial status report; (17) Bank release of guarantees; (18) Cash...
15 CFR 762.3 - Records exempt from recordkeeping requirements.
Code of Federal Regulations, 2011 CFR
2011-01-01
...; (2) Special export file list; (3) Vessel log from freight forwarder; (4) Inspection certificate; (5... form; (12) Financial hold form; (13) Export parts shipping problem form; (14) Draft number log; (15) Expense invoice mailing log; (16) Financial status report; (17) Bank release of guarantees; (18) Cash...
DOE Office of Scientific and Technical Information (OSTI.GOV)
Gunter, Dan; Lee, Jason; Stoufer, Martin
2003-03-28
The NetLogger Toolkit is designed to monitor, under actual operating conditions, the behavior of all the elements of the application-to-application communication path in order to determine exactly where time is spent within a complex system Using NetLogger, distnbuted application components are modified to produce timestamped logs of "interesting" events at all the critical points of the distributed system Events from each component are correlated, which allov^ one to characterize the performance of all aspects of the system and network in detail. The NetLogger Toolkit itself consists of four components an API and library of functions to simplify the generation ofmore » application-level event logs, a set of tools for collecting and sorting log files, an event archive system, and a tool for visualization and analysis of the log files In order to instrument an application to produce event logs, the application developer inserts calls to the NetLogger API at all the critical points in the code, then links the application with the NetLogger library All the tools in the NetLogger Toolkit share a common log format, and assume the existence of accurate and synchronized system clocks NetLogger messages can be logged using an easy-to-read text based format based on the lETF-proposed ULM format, or a binary format that can still be used through the same API but that is several times faster and smaller, with performance comparable or better than binary message formats such as MPI, XDR, SDDF-Binary, and PBIO. The NetLogger binary format is both highly efficient and self-describing, thus optimized for the dynamic message construction and parsing of application instrumentation. NetLogger includes an "activation" API that allows NetLogger logging to be turned on, off, or modified by changing an external file This IS useful for activating logging in daemons/services (e g GndFTP server). The NetLogger reliability API provides the ability to specify backup logging locations and penodically try to reconnect broken TCP pipe. A typical use for this is to store data on local disk while net is down. An event archiver can log one or more incoming NetLogger streams to a local disk file (netlogd) or to a mySQL database (netarchd). We have found exploratory, visual analysis of the log event data to be the most useful means of determining the causes of performance anomalies The NetLogger Visualization tool, niv, has been developed to provide a flexible and interactive graphical representation of system-level and application-level events.« less
Analysis of the request patterns to the NSSDC on-line archive
NASA Technical Reports Server (NTRS)
Johnson, Theodore
1994-01-01
NASA missions, both for earth science and for space science, collect huge amounts of data, and the rate at which data is being gathered is increasing. For example, the EOSDIS project is expected to collect petabytes per year. In addition, these archives are being made available to remote users over the Internet. The ability to manage the growth of the size and request activity of scientific archives depends on an understanding of the access patterns of scientific users. The National Space Science Data Center (NSSDC) of NASA Goddard Space Flight Center has run their on-line mass storage archive of space data, the National Data Archive and Distribution Service (NDADS), since November 1991. A large world-wide space research community makes use of NSSDC, requesting more than 20,000 files per month. Since the initiation of their service, they have maintained log files which record all accesses the archive. In this report, we present an analysis of the NDADS log files. We analyze the log files, and discuss several issues, including caching, reference patterns, clustering, and system loading.
Combining MLC and SVM Classifiers for Learning Based Decision Making: Analysis and Evaluations
Zhang, Yi; Ren, Jinchang; Jiang, Jianmin
2015-01-01
Maximum likelihood classifier (MLC) and support vector machines (SVM) are two commonly used approaches in machine learning. MLC is based on Bayesian theory in estimating parameters of a probabilistic model, whilst SVM is an optimization based nonparametric method in this context. Recently, it is found that SVM in some cases is equivalent to MLC in probabilistically modeling the learning process. In this paper, MLC and SVM are combined in learning and classification, which helps to yield probabilistic output for SVM and facilitate soft decision making. In total four groups of data are used for evaluations, covering sonar, vehicle, breast cancer, and DNA sequences. The data samples are characterized in terms of Gaussian/non-Gaussian distributed and balanced/unbalanced samples which are then further used for performance assessment in comparing the SVM and the combined SVM-MLC classifier. Interesting results are reported to indicate how the combined classifier may work under various conditions. PMID:26089862
Combining MLC and SVM Classifiers for Learning Based Decision Making: Analysis and Evaluations.
Zhang, Yi; Ren, Jinchang; Jiang, Jianmin
2015-01-01
Maximum likelihood classifier (MLC) and support vector machines (SVM) are two commonly used approaches in machine learning. MLC is based on Bayesian theory in estimating parameters of a probabilistic model, whilst SVM is an optimization based nonparametric method in this context. Recently, it is found that SVM in some cases is equivalent to MLC in probabilistically modeling the learning process. In this paper, MLC and SVM are combined in learning and classification, which helps to yield probabilistic output for SVM and facilitate soft decision making. In total four groups of data are used for evaluations, covering sonar, vehicle, breast cancer, and DNA sequences. The data samples are characterized in terms of Gaussian/non-Gaussian distributed and balanced/unbalanced samples which are then further used for performance assessment in comparing the SVM and the combined SVM-MLC classifier. Interesting results are reported to indicate how the combined classifier may work under various conditions.
Cooperative storage of shared files in a parallel computing system with dynamic block size
Bent, John M.; Faibish, Sorin; Grider, Gary
2015-11-10
Improved techniques are provided for parallel writing of data to a shared object in a parallel computing system. A method is provided for storing data generated by a plurality of parallel processes to a shared object in a parallel computing system. The method is performed by at least one of the processes and comprises: dynamically determining a block size for storing the data; exchanging a determined amount of the data with at least one additional process to achieve a block of the data having the dynamically determined block size; and writing the block of the data having the dynamically determined block size to a file system. The determined block size comprises, e.g., a total amount of the data to be stored divided by the number of parallel processes. The file system comprises, for example, a log structured virtual parallel file system, such as a Parallel Log-Structured File System (PLFS).
Federal Register 2010, 2011, 2012, 2013, 2014
2013-11-25
... to protest should file with the Federal Energy Regulatory Commission, 888 First Street NE... . To facilitate electronic service, persons with Internet access who will eFile a document and/or be...Registration link. Select the eFiling link to log on and submit the intervention or protests. Persons unable to...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-09-30
... to protest should file with the Federal Energy Regulatory Commission, 888 First Street NE... . To facilitate electronic service, persons with Internet access who will eFile a document and/or be...Registration link. Select the eFiling link to log on and submit the intervention or protests. Persons unable to...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-05-16
... link to log on and submit the intervention or protests. Persons unable to file electronically should... protest should file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-05-20
... link to log on and submit the intervention or protests. Persons unable to file electronically should... protest should file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2012-10-24
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-10-15
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-12-12
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-05-16
... link to log on and submit the intervention or protests. Persons unable to file electronically should... protest should file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-12-03
... link to log on and submit the intervention or protests. Persons unable to file electronically should... protest should file with the Federal Energy Regulatory Commission, 888 First Street, NE., Washington, DC... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-07-24
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-11-13
... to protest should file with the Federal Energy Regulatory Commission, 888 First Street NE... . To facilitate electronic service, persons with Internet access who will eFile a document and/or be...Registration link. Select the eFiling link to log on and submit the intervention or protests. Persons unable to...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-08-14
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2012-10-24
... intervene or to protest should file with the Federal Energy Regulatory Commission, 888 First Street NE...://www.ferc.gov . To facilitate electronic service, persons with Internet access who will eFile a... using the eRegistration link. Select the eFiling link to log on and submit the intervention or protests...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-08-02
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-05-16
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-09-25
... to protest should file with the Federal Energy Regulatory Commission, 888 First Street, NE... . To facilitate electronic service, persons with Internet access who will eFile a document and/or be...Registration link. Select the eFiling link to log on and submit the intervention or protests. Persons unable to...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-12-12
... link to log on and submit the intervention or protests. Persons unable to file electronically should... protest should file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2012-10-24
... link to log on and submit the intervention or protests. Persons unable to file electronically should... protest should file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2012-10-24
... to protest should file with the Federal Energy Regulatory Commission, 888 First Street NE... . To facilitate electronic service, persons with Internet access who will eFile a document and/or be...Registration link. Select the eFiling link to log on and submit the intervention or protests. Persons unable to...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-10-15
... protest should file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC... . To facilitate electronic service, persons with Internet access who will eFile a document and/or be...Registration link. Select the eFiling link to log on and submit the intervention or protests. Persons unable to...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-08-27
... to protest should file with the Federal Energy Regulatory Commission, 888 First Street NE... . To facilitate electronic service, persons with Internet access who will eFile a document and/or be...Registration link. Select the eFiling link to log on and submit the intervention or protests. Persons unable to...
SedMob: A mobile application for creating sedimentary logs in the field
NASA Astrophysics Data System (ADS)
Wolniewicz, Pawel
2014-05-01
SedMob is an open-source, mobile software package for creating sedimentary logs, targeted for use in tablets and smartphones. The user can create an unlimited number of logs, save data from each bed in the log as well as export and synchronize the data with a remote server. SedMob is designed as a mobile interface to SedLog: a free multiplatform package for drawing graphic logs that runs on PC computers. Data entered into SedMob are saved in the CSV file format, fully compatible with SedLog.
COMBATXXI, JDAFS, and LBC Integration Requirements for EASE
2015-10-06
process as linear and as new data is made available, any previous analysis is obsolete and has to start the process over again. Figure 2 proposes a...final line of the manifest file names the scenario file associated with the run. Under the usual practice, the analyst now starts the COMBATXXI...describes which events are to be logged. Finally the scenario is started with the click of a button. The simulation generates logs of a couple of sorts
Model Analyst’s Toolkit User Guide, Version 7.1.0
2015-08-01
Help > About) Environment details ( operating system ) metronome.log file, located in your MAT 7.1.0 installation folder Any log file that...requirements to run the Model Analyst’s Toolkit: Windows XP operating system (or higher) with Service Pack 2 and all critical Windows updates installed...application icon on your desktop Create a Quick Launch icon – Creates a MAT application icon on the taskbar for operating systems released
Users' information-seeking behavior on a medical library Website
Rozic-Hristovski, Anamarija; Hristovski, Dimitar; Todorovski, Ljupco
2002-01-01
The Central Medical Library (CMK) at the Faculty of Medicine, University of Ljubljana, Slovenia, started to build a library Website that included a guide to library services and resources in 1997. The evaluation of Website usage plays an important role in its maintenance and development. Analyzing and exploring regularities in the visitors' behavior can be used to enhance the quality and facilitate delivery of information services, identify visitors' interests, and improve the server's performance. The analysis of the CMK Website users' navigational behavior was carried out by analyzing the Web server log files. These files contained information on all user accesses to the Website and provided a great opportunity to learn more about the behavior of visitors to the Website. The majority of the available tools for Web log file analysis provide a predefined set of reports showing the access count and the transferred bytes grouped along several dimensions. In addition to the reports mentioned above, the authors wanted to be able to perform interactive exploration and ad hoc analysis and discover trends in a user-friendly way. Because of that, we developed our own solution for exploring and analyzing the Web logs based on data warehousing and online analytical processing technologies. The analytical solution we developed proved successful, so it may find further application in the field of Web log file analysis. We will apply the findings of the analysis to restructuring the CMK Website. PMID:11999179
DOE Office of Scientific and Technical Information (OSTI.GOV)
Hwang, M; Jang, S; Ozhasoglu, C
2015-06-15
Purpose: The InCise™ Multileaf Collimator (MLC) of CyberKnife M6™ System has been released recently. The purpose of this study was to explore the dosimetric characteristics of the new MLC. In particular, the penumbra characteristics of MLC fields at varying locations are evaluated. Methods: EBT3-based film measurements were performed with varying MLC fields ranging from 7.5 mm to 27.5 mm. Seventeen regions of interests (ROIs) were identified for irradiation. These are regions located at the central area (denoted as reference field), at the left/right edge areas of reference open field, at an intermediate location between central and edge area. Single beammore » treatment plans were designed by using the MultiPlan and was delivered using the Blue Phantom. Gafchromic films were irradiated at 1.5 cm depth in the Blue Phantom and analyzed using the Film Pro software. Variation of maximum dose, penumbra of MLC-defined fields, and symmetry/flatness were calculated as a function of locations of MLC fields. Results: The InCise™ MLC System showed relatively consistent dose distribution and penumbra size with varying locations of MLC fields. The measured maximum dose varied within 5 % at different locations compared to that at the central location and agreed with the calculated data well within 2%. The measured penumbrae were in the range of 2.9 mm and 3.7 mm and were relatively consistent regardless of locations. However, dose profiles in the out-of-field and in-field regions varied with locations and field sizes. Strong variation was seen for all fields located at 55 mm away from the central field. The MLC leakage map showed that the leakage is dependent on position. Conclusion: The size of penumbra and normalized maximum dose for MLC-defined fields were consistent in different regions of MLC. However, dose profiles in the out-field region varied with locations and field sizes.« less
SU-E-T-293: Simplifying Assumption for Determining Sc and Sp
DOE Office of Scientific and Technical Information (OSTI.GOV)
King, R; Cheung, A; Anderson, R
Purpose: Scp(mlc,jaw) is a two-dimensional function of collimator field size and effective field size. Conventionally, Scp(mlc,jaw) is treated as separable into components Sc(jaw) and Sp(mlc). Scp(mlc=jaw) is measured in phantom and Sc(jaw) is measured in air with Sp=Scp/Sc. Ideally, Sc and Sp would be able to predict measured values of Scp(mlc,jaw) for all combinations of mlc and jaw. However, ideal Sc and Sp functions do not exist and a measured two-dimensional Scp dataset cannot be decomposed into a unique pair of one-dimensional functions.If the output functions Sc(jaw) and Sp(mlc) were equal to each other and thus each equal to Scp(mlc=jaw){supmore » 0.5}, this condition would lead to a simpler measurement process by eliminating the need for in-air measurements. Without the distorting effect of the buildup-cap, small-field measurement would be limited only by the dimensions of the detector and would thus be improved by this simplification of the output functions. The goal of the present study is to evaluate an assumption that Sc=Sp. Methods: For a 6 MV x-ray beam, Sc and Sp were determined both by the conventional method and as Scp(mlc=jaw){sup 0.5}. Square field benchmark values of Scp(mlc,jaw) were then measured across the range from 2×2 to 29×29. Both Sc and Sp functions were then evaluated as to their ability to predict these measurements. Results: Both methods produced qualitatively similar results with <4% error for all cases and >3% error in 1 case. The conventional method produced 2 cases with >2% error, while the squareroot method produced only 1 such case. Conclusion: Though it would need to be validated for any specific beam to which it might be applied, under the conditions studied, the simplifying assumption that Sc = Sp is justified.« less
18 CFR 270.304 - Tight formation gas.
Code of Federal Regulations, 2011 CFR
2011-04-01
... determination that natural gas is tight formation gas must file with the jurisdictional agency an application... formation; (d) A complete copy of the well log, including the log heading identifying the designated tight...
SU-F-T-177: Impacts of Gantry Angle Dependent Scanning Beam Properties for Proton Treatment
DOE Office of Scientific and Technical Information (OSTI.GOV)
Lin, Y; Clasie, B; Lu, H
Purpose: In pencil beam scanning (PBS), the delivered spot MU, position and size are slightly different at different gantry angles. We investigated the level of delivery uncertainty at different gantry angles through a log file analysis. Methods: 34 PBS fields covering full 360 degrees gantry angle spread were collected retrospectively from 28 patients treated at our institution. All fields were delivered at zero gantry angle and the prescribed gantry angle, and measured at isocenter with the MatriXX 2D array detector at the prescribed gantry angle. The machine log files were analyzed to extract the delivered MU per spot and themore » beam position from the strip ionization chambers in the treatment nozzle. The beam size was separately measured as a function of gantry angle and beam energy. Using this information, the dose was calculated in a water phantom at both gantry angles and compared to the measurement using the 3D γ-index at 2mm/2%. Results: The spot-by-spot difference between the beam position in the log files from the delivery at the two gantry angles has a mean of 0.3 and 0.4 mm and a standard deviation of 0.6 and 0.7 mm for × and y directions, respectively. Similarly, the spot-by-spot difference between the MU in the log files from the delivery at the two gantry angles has a mean 0.01% and a standard deviation of 0.7%. These small deviations lead to an excellent agreement in dose calculations with an average γ pass rate for all fields being approximately 99.7%. When each calculation is compared to the measurement, a high correlation in γ was also found. Conclusion: Using machine logs files, we verified that PBS beam delivery at different gantry angles are sufficiently small and the planned spot position and MU. This study brings us one step closer to simplifying our patient-specific QA.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Xiong, Z; Vijayan, S; Rana, V
2015-06-15
Purpose: A system was developed that automatically calculates the organ and effective dose for individual fluoroscopically-guided procedures using a log of the clinical exposure parameters. Methods: We have previously developed a dose tracking system (DTS) to provide a real-time color-coded 3D- mapping of skin dose. This software produces a log file of all geometry and exposure parameters for every x-ray pulse during a procedure. The data in the log files is input into PCXMC, a Monte Carlo program that calculates organ and effective dose for projections and exposure parameters set by the user. We developed a MATLAB program to readmore » data from the log files produced by the DTS and to automatically generate the definition files in the format used by PCXMC. The processing is done at the end of a procedure after all exposures are completed. Since there are thousands of exposure pulses with various parameters for fluoroscopy, DA and DSA and at various projections, the data for exposures with similar parameters is grouped prior to entry into PCXMC to reduce the number of Monte Carlo calculations that need to be performed. Results: The software developed automatically transfers data from the DTS log file to PCXMC and runs the program for each grouping of exposure pulses. When the dose from all exposure events are calculated, the doses for each organ and all effective doses are summed to obtain procedure totals. For a complicated interventional procedure, the calculations can be completed on a PC without manual intervention in less than 30 minutes depending on the level of data grouping. Conclusion: This system allows organ dose to be calculated for individual procedures for every patient without tedious calculations or data entry so that estimates of stochastic risk can be obtained in addition to the deterministic risk estimate provided by the DTS. Partial support from NIH grant R01EB002873 and Toshiba Medical Systems Corp.« less
The first clinical implementation of electromagnetic transponder-guided MLC tracking.
Keall, Paul J; Colvill, Emma; O'Brien, Ricky; Ng, Jin Aun; Poulsen, Per Rugaard; Eade, Thomas; Kneebone, Andrew; Booth, Jeremy T
2014-02-01
We report on the clinical process, quality assurance, and geometric and dosimetric results of the first clinical implementation of electromagnetic transponder-guided MLC tracking which occurred on 28 November 2013 at the Northern Sydney Cancer Centre. An electromagnetic transponder-based positioning system (Calypso) was modified to send the target position output to in-house-developed MLC tracking code, which adjusts the leaf positions to optimally align the treatment beam with the real-time target position. Clinical process and quality assurance procedures were developed and performed. The first clinical implementation of electromagnetic transponder-guided MLC tracking was for a prostate cancer patient being treated with dual-arc VMAT (RapidArc). For the first fraction of the first patient treatment of electromagnetic transponder-guided MLC tracking we recorded the in-room time and transponder positions, and performed dose reconstruction to estimate the delivered dose and also the dose received had MLC tracking not been used. The total in-room time was 21 min with 2 min of beam delivery. No additional time was needed for MLC tracking and there were no beam holds. The average prostate position from the initial setup was 1.2 mm, mostly an anterior shift. Dose reconstruction analysis of the delivered dose with MLC tracking showed similar isodose and target dose volume histograms to the planned treatment and a 4.6% increase in the fractional rectal V60. Dose reconstruction without motion compensation showed a 30% increase in the fractional rectal V60 from that planned, even for the small motion. The real-time beam-target correction method, electromagnetic transponder-guided MLC tracking, has been translated to the clinic. This achievement represents a milestone in improving geometric and dosimetric accuracy, and by inference treatment outcomes, in cancer radiotherapy.
The first clinical implementation of electromagnetic transponder-guided MLC tracking
Keall, Paul J.; Colvill, Emma; O’Brien, Ricky; Ng, Jin Aun; Poulsen, Per Rugaard; Eade, Thomas; Kneebone, Andrew; Booth, Jeremy T.
2014-01-01
Purpose: We report on the clinical process, quality assurance, and geometric and dosimetric results of the first clinical implementation of electromagnetic transponder-guided MLC tracking which occurred on 28 November 2013 at the Northern Sydney Cancer Centre. Methods: An electromagnetic transponder-based positioning system (Calypso) was modified to send the target position output to in-house-developed MLC tracking code, which adjusts the leaf positions to optimally align the treatment beam with the real-time target position. Clinical process and quality assurance procedures were developed and performed. The first clinical implementation of electromagnetic transponder-guided MLC tracking was for a prostate cancer patient being treated with dual-arc VMAT (RapidArc). For the first fraction of the first patient treatment of electromagnetic transponder-guided MLC tracking we recorded the in-room time and transponder positions, and performed dose reconstruction to estimate the delivered dose and also the dose received had MLC tracking not been used. Results: The total in-room time was 21 min with 2 min of beam delivery. No additional time was needed for MLC tracking and there were no beam holds. The average prostate position from the initial setup was 1.2 mm, mostly an anterior shift. Dose reconstruction analysis of the delivered dose with MLC tracking showed similar isodose and target dose volume histograms to the planned treatment and a 4.6% increase in the fractional rectal V60. Dose reconstruction without motion compensation showed a 30% increase in the fractional rectal V60 from that planned, even for the small motion. Conclusions: The real-time beam-target correction method, electromagnetic transponder-guided MLC tracking, has been translated to the clinic. This achievement represents a milestone in improving geometric and dosimetric accuracy, and by inference treatment outcomes, in cancer radiotherapy. PMID:24506591
SU-E-T-430: Modeling MLC Leaf End in 2D for Sliding Window IMRT and Arc Therapy
DOE Office of Scientific and Technical Information (OSTI.GOV)
Liang, X; Zhu, T
2014-06-01
Purpose: To develop a 2D geometric model for MLC accounting for leaf end dose leakage for dynamic IMRT and Rapidarc therapy. Methods: Leaf-end dose leakage is one of the problems for MLC dose calculation and modeling. Dosimetric leaf gap used to model the MLC and to count for leakage in dose calculation, but may not be accurate for smaller leaf gaps. We propose another geometric modeling method to compensate for the MLC round-shape leaf ends dose leakage, and improve the accuracy of dose calculation and dose verification. A triangular function is used to geometrically model the MLC leaf end leakagemore » in the leaf motion direction, and a step function is used in the perpendicular direction. Dose measurements with different leaf gap, different window width, and different window height were conducted, and the results were used to fit the analytical model to get the model parameters. Results: Analytical models have been obtained for stop-and-shoot and dynamic modes for MLC motion. Parameters a=0.4, lw'=5.0 mm for 6X and a=0.54, lw'=4.1 mm for 15x were obtained from the fitting process. The proposed MLC leaf end model improves the dose profile at the two ends of the sliding window opening. This improvement is especially significant for smaller sliding window openings, which are commonly used for highly modulated IMRT plans and arc therapy plans. Conclusion: This work models the MLC round leaf end shape and movement pattern for IMRT dose calculation. The theory, as well as the results in this work provides a useful tool for photon beam IMRT dose calculation and verification.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Götstedt, Julia; Karlsson Hauer, Anna; Bäck, Anna, E-mail: anna.back@vgregion.se
Purpose: Complexity metrics have been suggested as a complement to measurement-based quality assurance for intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT). However, these metrics have not yet been sufficiently validated. This study develops and evaluates new aperture-based complexity metrics in the context of static multileaf collimator (MLC) openings and compares them to previously published metrics. Methods: This study develops the converted aperture metric and the edge area metric. The converted aperture metric is based on small and irregular parts within the MLC opening that are quantified as measured distances between MLC leaves. The edge area metricmore » is based on the relative size of the region around the edges defined by the MLC. Another metric suggested in this study is the circumference/area ratio. Earlier defined aperture-based complexity metrics—the modulation complexity score, the edge metric, the ratio monitor units (MU)/Gy, the aperture area, and the aperture irregularity—are compared to the newly proposed metrics. A set of small and irregular static MLC openings are created which simulate individual IMRT/VMAT control points of various complexities. These are measured with both an amorphous silicon electronic portal imaging device and EBT3 film. The differences between calculated and measured dose distributions are evaluated using a pixel-by-pixel comparison with two global dose difference criteria of 3% and 5%. The extent of the dose differences, expressed in terms of pass rate, is used as a measure of the complexity of the MLC openings and used for the evaluation of the metrics compared in this study. The different complexity scores are calculated for each created static MLC opening. The correlation between the calculated complexity scores and the extent of the dose differences (pass rate) are analyzed in scatter plots and using Pearson’s r-values. Results: The complexity scores calculated by the edge area metric, converted aperture metric, circumference/area ratio, edge metric, and MU/Gy ratio show good linear correlation to the complexity of the MLC openings, expressed as the 5% dose difference pass rate, with Pearson’s r-values of −0.94, −0.88, −0.84, −0.89, and −0.82, respectively. The overall trends for the 3% and 5% dose difference evaluations are similar. Conclusions: New complexity metrics are developed. The calculated scores correlate to the complexity of the created static MLC openings. The complexity of the MLC opening is dependent on the penumbra region relative to the area of the opening. The aperture-based complexity metrics that combined either the distances between the MLC leaves or the MLC opening circumference with the aperture area show the best correlation with the complexity of the static MLC openings.« less
Federal Register 2010, 2011, 2012, 2013, 2014
2013-05-16
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-05-16
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-05-16
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-12-12
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2012-10-24
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2012-11-21
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2012-10-24
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2012-08-31
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-09-30
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-10-07
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2012-11-21
... link to log on and submit the intervention or protests. Persons unable to file electronically should... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-12-12
... file with the Federal Energy Regulatory Commission, 888 First Street NE., Washington, DC 20426, in... . To facilitate electronic service, persons with Internet access who will eFile a document and/or be...Registration link. Select the eFiling link to log on and submit the intervention or protests. Persons unable to...
Myosin Light Chain Kinase Mediates Intestinal Barrier Disruption following Burn Injury
Chen, Chuanli; Wang, Pei; Su, Qin; Wang, Shiliang; Wang, Fengjun
2012-01-01
Background Severe burn injury results in the loss of intestinal barrier function, however, the underlying mechanism remains unclear. Myosin light chain (MLC) phosphorylation mediated by MLC kinase (MLCK) is critical to the pathophysiological regulation of intestinal barrier function. We hypothesized that the MLCK-dependent MLC phosphorylation mediates the regulation of intestinal barrier function following burn injury, and that MLCK inhibition attenuates the burn-induced intestinal barrier disfunction. Methodology/Principal Findings Male balb/c mice were assigned randomly to either sham burn (control) or 30% total body surface area (TBSA) full thickness burn without or with intraperitoneal injection of ML-9 (2 mg/kg), an MLCK inhibitor. In vivo intestinal permeability to fluorescein isothiocyanate (FITC)-dextran was measured. Intestinal mucosa injury was assessed histologically. Tight junction proteins ZO-1, occludin and claudin-1 was analyzed by immunofluorescent assay. Expression of MLCK and phosphorylated MLC in ileal mucosa was assessed by Western blot. Intestinal permeability was increased significantly after burn injury, which was accompanied by mucosa injury, tight junction protein alterations, and increase of both MLCK and MLC phosphorylation. Treatment with ML-9 attenuated the burn-caused increase of intestinal permeability, mucosa injury, tight junction protein alterations, and decreased MLC phosphorylation, but not MLCK expression. Conclusions/Significance The MLCK-dependent MLC phosphorylation mediates intestinal epithelial barrier dysfunction after severe burn injury. It is suggested that MLCK-dependent MLC phosphorylation may be a critical target for the therapeutic treatment of intestinal epithelial barrier disruption after severe burn injury. PMID:22529961
Keall, Paul J; Colvill, Emma; O'Brien, Ricky; Caillet, Vincent; Eade, Thomas; Kneebone, Andrew; Hruby, George; Poulsen, Per R; Zwan, Benjamin; Greer, Peter B; Booth, Jeremy
2018-06-01
To report on the primary and secondary outcomes of a prospective clinical trial of electromagnetic-guided multileaf collimator (MLC) tracking radiation therapy for prostate cancer. Twenty-eight men with prostate cancer were treated with electromagnetic-guided MLC tracking with volumetric modulated arc therapy. A total of 858 fractions were delivered, with the dose per fraction ranging from 2 to 13.75 Gy. The primary outcome was feasibility, with success determined if >95% of fractions were successfully delivered. The secondary outcomes were (1) the improvement in beam-target geometric alignment, (2) the improvement in dosimetric coverage of the prostate and avoidance of critical structures, and (3) no acute grade ≥3 genitourinary or gastrointestinal toxicity. All 858 planned fractions were successfully delivered with MLC tracking, demonstrating the primary outcome of feasibility (P < .001). MLC tracking improved the beam-target geometric alignment from 1.4 to 0.90 mm (root-mean-square error). MLC tracking improved the dosimetric coverage of the prostate and reduced the daily variation in dose to critical structures. No acute grade ≥3 genitourinary or gastrointestinal toxicity was observed. Electromagnetic-guided MLC tracking radiation therapy for prostate cancer is feasible. The patients received improved geometric targeting and delivered dose distributions that were closer to those planned than they would have received without electromagnetic-guided MLC tracking. No significant acute toxicity was observed. Copyright © 2018 Elsevier Inc. All rights reserved.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Thind, K; Tolakanahalli, R
2014-08-15
The aim of this study was to analyze the feasibility of designing comprehensive QA plans using iComCAT for Elekta machines equipped with Agility multileaf collimator and continuously variable dose rate. Test plans with varying MLC speed, gantry speed, and dose rate were created and delivered in a controlled manner. A strip test was designed with three 1 cm MLC positions and delivered using dynamic, StepNShoot and VMAT techniques. Plans were also designed to test error in MLC position with various gantry speeds and various MLC speeds. The delivery fluence was captured using the electronic portal-imaging device. Gantry speed was foundmore » to be within tolerance as per the Canadian standards. MLC positioning errors at higher MLC speed with gravity effects does add more than 2 mm discrepancy. More tests need to be performed to evaluate MLC performance using independent measurement systems. The treatment planning system with end-to-end testing necessary for commissioning was also investigated and found to have >95% passing rates within 3%/3mm gamma criteria. Future studies involve performing off-axis gantry starshot pattern and repeating the tests on three matched Elekta linear accelerators.« less
Spatial variation of dosimetric leaf gap and its impact on dose delivery
DOE Office of Scientific and Technical Information (OSTI.GOV)
Kumaraswamy, Lalith K., E-mail: Lalith.Kumaraswamy@roswellpark.org; Schmitt, Jonathan D.; Bailey, Daniel W.
Purpose: During dose calculation, the Eclipse treatment planning system (TPS) retracts the multileaf collimator (MLC) leaf positions by half of the dosimetric leaf gap (DLG) value (measured at central axis) for all leaf positions in a dynamic MLC plan to accurately model the rounded leaf ends. The aim of this study is to map the variation of DLG along the travel path of each MLC leaf pair and quantify how this variation impacts delivered dose. Methods: 6 MV DLG values were measured for all MLC leaf pairs in increments of 1.0 cm (from the line intersecting the CAX and perpendicularmore » to MLC motion) to 13.0 cm off axis distance at dmax. The measurements were performed on two Varian linear accelerators, both employing the Millennium 120-leaf MLCs. The measurements were performed at several locations in the beam with both a Sun Nuclear MapCHECK device and a PTW pinpoint ion chamber. Results: The measured DLGs for the middle 40 MLC leaf pairs (each 0.5 cm width) at positions along a line through the CAX and perpendicular to MLC leaf travel direction were very similar, varying maximally by only 0.2 mm. The outer 20 MLC leaf pairs (each 1.0 cm width) have much lower DLG values, about 0.3–0.5 mm lower than the central MLC leaf pair, at their respective central line position. Overall, the mean and the maximum variation between the 0.5 cm width leaves and the 1.0 cm width leaf pairs are 0.32 and 0.65 mm, respectively. Conclusions: The spatial variation in DLG is caused by the variation of intraleaf transmission through MLC leaves. Fluences centered on the CAX would not be affected since DLG does not vary; but any fluences residing significantly off axis with narrow sweeping leaves may exhibit significant dose differences. This is due to the fact that there are differences in DLG between the true DLG exhibited by the 1.0 cm width outer leaves and the constant DLG value utilized by the TPS for dose calculation. Since there are large differences in DLG between the 0.5 cm width leaf pairs and 1.0 cm width leaf pairs, there is a need to correct the TPS plans, especially those with high modulation (narrow dynamic MLC gap), with 2D variation of DLG.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Shang Qingyang; Qi Peng; Ferjani, Samah
2013-06-15
Purpose: The aim of the study was to evaluate the impact of multileaf collimator (MLC) leaf width on treatment adaptation and delivery accuracy for concurrent treatment of the prostate and pelvic lymph nodes with intensity modulated radiation therapy (IMRT). Methods: Seventy-five kilovoltage cone beam CTs (KV-CBCT) from six patients were included for this retrospective study. For each patient, three different IMRT plans were created based on a planning CT using three different MLC leaf widths of 2.5, 5, and 10 mm, respectively. For each CBCT, the prostate displacement was determined by a dual image registration. Adaptive plans were created bymore » shifting selected MLC leaf pairs to compensate for daily prostate movements. To evaluate the impact of MLC leaf width on the adaptive plan for each daily CBCT, three MLC shifted plans were created using three different leaf widths of MLCs (a total of 225 adaptive treatment plans). Selective dosimetric endpoints for the tumor volumes and organs at risk (OARs) were evaluated for these adaptive plans. Using the planning CT from a selected patient, MLC shifted plans for three hypothetical longitudinal shifts of 2, 4, and 8 mm were delivered on the three linear accelerators to test the deliverability of the shifted plans and to compare the dose accuracy of the shifted plans with the original IMRT plans. Results: Adaptive plans from 2.5 and 5 mm MLCs had inadequate dose coverage to the prostate (D99 < 97%, or D{sub mean} < 99% of the planned dose) in 6%-8% of the fractions, while adaptive plans from 10 mm MLC led to inadequate dose coverage to the prostate in 25.3% of the fractions. The average V{sub 56Gy} of the prostate over the six patients was improved by 6.4% (1.6%-32.7%) and 5.8% (1.5%-35.7%) with adaptive plans from 2.5 and 5 mm MLCs, respectively, when compared with adaptive plans from 10 mm MLC. Pelvic lymph nodes were well covered for all MLC adaptive plans, as small differences were observed for D99, D{sub mean}, and V{sub 50.4Gy}. Similar OAR sparing could be achieved for the bladder and rectum with all three MLCs for treatment adaptation. The MLC shifted plans can be accurately delivered on all three linear accelerators with accuracy similar to their original IMRT plans, where gamma (3%/3 mm) passing rates were 99.6%, 93.0%, and 92.1% for 2.5, 5, and 10 mm MLCs, respectively. The percentages of pixels with dose differences between the measurement and calculation being less than 3% of the maximum dose were 85.9%, 82.5%, and 70.5% for the original IMRT plans from the three MLCs, respectively. Conclusions: Dosimetric advantages associated with smaller MLC leaves were observed in terms of the coverage to the prostate, when the treatment was adapted to account for daily prostate movement for concurrent irradiation of the prostate and pelvic lymph nodes. The benefit of switching the MLC from 10 to 5 mm was significant (p Much-Less-Than 0.01); however, switching the MLC from 5 to 2.5 mm would not gain significant (p= 0.15) improvement. IMRT plans with smaller MLC leaf widths achieved more accurate dose delivery.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Bronevetsky, G.
2014-09-01
Enables applications to emit log information into an output file and produced a structured visual summary of the log data, as well as various statistical analyses of it. This makes it easier for developers to understand the behavior of their applications.
Federal Register 2010, 2011, 2012, 2013, 2014
2010-09-29
... the respondents, including the use of automated collection techniques or other forms of information...: OMB Control Number: 3060-0360. Title: Section 80.409, Station Logs. Form No.: N/A. Type of Review... for filing suits upon such claims. Section 80.409(d), Ship Radiotelegraph Logs: Logs of ship stations...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-05-16
... assumptions of liability. Any person desiring to intervene or to protest should file with the Federal Energy... access who will eFile a document and/or be listed as a contact for an intervenor must create and validate an eRegistration account using the eRegistration link. Select the eFiling link to log on and submit...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-08-14
... securities and assumptions of liability. Any person desiring to intervene or to protest should file with the... with Internet access who will eFile a document and/or be listed as a contact for an intervenor must create and validate an eRegistration account using the eRegistration link. Select the eFiling link to log...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-08-14
... of liability. Any person desiring to intervene or to protest should file with the Federal Energy... access who will eFile a document and/or be listed as a contact for an intervenor must create and validate an eRegistration account using the eRegistration link. Select the eFiling link to log on and submit...
Federal Register 2010, 2011, 2012, 2013, 2014
2012-10-24
... assumptions of liability. Any person desiring to intervene or to protest should file with the Federal Energy... Internet access who will eFile a document and/or be listed as a contact for an intervenor must create and validate an eRegistration account using the eRegistration link. Select the eFiling link to log on and...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-07-05
... assumptions of liability. Any person desiring to intervene or to protest should file with the Federal Energy... access who will eFile a document and/or be listed as a contact for an intervenor must create and validate an eRegistration account using the eRegistration link. Select the eFiling link to log on and submit...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-05-16
... assumptions of liability. Any person desiring to intervene or to protest should file with the Federal Energy... access who will eFile a document and/or be listed as a contact for an intervenor must create and validate an eRegistration account using the eRegistration link. Select the eFiling link to log on and submit...
SU-E-T-613: Dosimetric Consequences of Systematic MLC Leaf Positioning Errors
DOE Office of Scientific and Technical Information (OSTI.GOV)
Kathuria, K; Siebers, J
2014-06-01
Purpose: The purpose of this study is to determine the dosimetric consequences of systematic MLC leaf positioning errors for clinical IMRT patient plans so as to establish detection tolerances for quality assurance programs. Materials and Methods: Dosimetric consequences were simulated by extracting mlc delivery instructions from the TPS, altering the file by the specified error, reloading the delivery instructions into the TPS, recomputing dose, and extracting dose-volume metrics for one head-andneck and one prostate patient. Machine error was simulated by offsetting MLC leaves in Pinnacle in a systematic way. Three different algorithms were followed for these systematic offsets, and aremore » as follows: a systematic sequential one-leaf offset (one leaf offset in one segment per beam), a systematic uniform one-leaf offset (same one leaf offset per segment per beam) and a systematic offset of a given number of leaves picked uniformly at random from a given number of segments (5 out of 10 total). Dose to the PTV and normal tissue was simulated. Results: A systematic 5 mm offset of 1 leaf for all delivery segments of all beams resulted in a maximum PTV D98 deviation of 1%. Results showed very low dose error in all reasonably possible machine configurations, rare or otherwise, which could be simulated. Very low error in dose to PTV and OARs was shown in all possible cases of one leaf per beam per segment being offset (<1%), or that of only one leaf per beam being offset (<.2%). The errors resulting from a high number of adjacent leaves (maximum of 5 out of 60 total leaf-pairs) being simultaneously offset in many (5) of the control points (total 10–18 in all beams) per beam, in both the PTV and the OARs analyzed, were similarly low (<2–3%). Conclusions: The above results show that patient shifts and anatomical changes are the main source of errors in dose delivered, not machine delivery. These two sources of error are “visually complementary” and uncorrelated (albeit not additive in the final error) and one can easily incorporate error resulting from machine delivery in an error model based purely on tumor motion.« less
Liu, Shi; Wu, Yu; Wooten, H. Omar; Green, Olga; Archer, Brent; Li, Harold
2016-01-01
A software tool is developed, given a new treatment plan, to predict treatment delivery time for radiation therapy (RT) treatments of patients on ViewRay magnetic resonance image‐guided radiation therapy (MR‐IGRT) delivery system. This tool is necessary for managing patient treatment scheduling in our clinic. The predicted treatment delivery time and the assessment of plan complexities could also be useful to aid treatment planning. A patient's total treatment delivery time, not including time required for localization, is modeled as the sum of four components: 1) the treatment initialization time; 2) the total beam‐on time; 3) the gantry rotation time; and 4) the multileaf collimator (MLC) motion time. Each of the four components is predicted separately. The total beam‐on time can be calculated using both the planned beam‐on time and the decay‐corrected dose rate. To predict the remain‐ing components, we retrospectively analyzed the patient treatment delivery record files. The initialization time is demonstrated to be random since it depends on the final gantry angle of the previous treatment. Based on modeling the relationships between the gantry rotation angles and the corresponding rotation time, linear regression is applied to predict the gantry rotation time. The MLC motion time is calculated using the leaves delay modeling method and the leaf motion speed. A quantitative analysis was performed to understand the correlation between the total treatment time and the plan complexity. The proposed algorithm is able to predict the ViewRay treatment delivery time with the average prediction error 0.22 min or 1.82%, and the maximal prediction error 0.89 min or 7.88%. The analysis has shown the correlation between the plan modulation (PM) factor and the total treatment delivery time, as well as the treatment delivery duty cycle. A possibility has been identified to significantly reduce MLC motion time by optimizing the positions of closed MLC pairs. The accuracy of the proposed prediction algorithm is sufficient to support patient treatment appointment scheduling. This developed software tool is currently applied in use on a daily basis in our clinic, and could also be used as an important indicator for treatment plan complexity. PACS number(s): 87.55.N PMID:27074472
Comparision of beam data requirements for MLC commissioning on a TPS
NASA Astrophysics Data System (ADS)
Solaiappan, Gopi; Singaravelu, Ganesan; Prakasarao, Aruna; Supe, Sanjay S.
2008-01-01
The treatment planning system (TPS) has become a key element in the radiotherapy process with the introduction of computer tomography (CT) based 3D conformal treatment planning. Commissioning of a MLC on a TPS either for conformal radiotherapy or intensity modulated radiation therapy (IMRT) requires beam data to be generated on a linear accelerator. Most of the TPS require these beam data to be generated with routine collimator jaws. However some TPS demand the data to be provided for MLC shaped fields. This prompted us to investigate whether beam data with jaws differ than that with MLC and whether the jaw based beam data would suffice for the commissioning of a MLC on a TPS. Beam data like percentage depth dose (PDD), cross beam profiles and output factors was acquired for jaws and MLC defined square fields for 6, 10 and 23 MV photon beams. Percentage depth dose and cross beam profiles were acquired with a radiation field analyzer RFA-200, CC13-S ion chambers with active volume of 0.13 cm3 and OmniPro-Accept software from Scanditronix-Wellhofer. A Medtec-TG51 water tank with Max-4000 electrometer and 0.6 cc PTW ionization chamber and a mini phantom from Standard Imaging was utilized for output measurements for millennium-120 MLC (Varian Medical Systems) and SRS diode detector (Scanditronix-Wellhofer) of 0.6 mm diameter of active area and 0.3 mm of active volume thickness for micro-MLC (BrainLab). The difference in PDD in the build-up region for millennium MLC was ±1.0% for 6 MV photons. For 10 MV photons the PDD difference was within ±4.0%. The difference in PDD for 23 MV photons ranged from 0% to 40.0%. PDD difference from build-up depth to about 28 cm was within ±1.0%. Difference in PDD crossed ±1.0% at 30 cm depth for 6 MV photons. The difference in PDD in the build-up region for mMLC was ±8.0% for 6 MV photons. For the smallest field size studied with micro-MLC i.e. 0.6 × 0.6 cm2 difference in PDD was more than ±1.0% in the build-up region and beyond a depth of 8.0 cm. The profiles for jaws and MLC agreed within the umbra region. However in the penumbra region small differences in doses were observed. The collimator scatter factor (Sc), phantom scatter factor (Sp) and output factor values for MLC were different that those for jaws. The differences in beam characteristics could have implication for intensity modulated radiation therapy and stereotactic radiosurgery in terms of dose in the build up region, exit dose, dose to the planning target volume (PTV) and organ at risk (OAR). Impact of these dosimetric differences between jaw and MLC needs to be further studied in terms of dose volume histograms for PTV and OAR and its further impact on tumor control probability (TCP) and normal tissue complication probability (NTCP).
17 CFR 259.602 - Form ID, uniform application for access codes to file on EDGAR.
Code of Federal Regulations, 2011 CFR
2011-04-01
...)—allows a filer, filing agent or training agent to log on to the EDGAR system, submit filings, and change... agent to change its Password. [69 FR 22710, Apr. 26, 2004] Editorial Note: For Federal Register... section of the printed volume and at at www.fdsys.gov. ...
17 CFR 259.602 - Form ID, uniform application for access codes to file on EDGAR.
Code of Federal Regulations, 2010 CFR
2010-04-01
...)—allows a filer, filing agent or training agent to log on to the EDGAR system, submit filings, and change... agent to change its Password. [69 FR 22710, Apr. 26, 2004] Editorial Note: For Federal Register... section of the printed volume and on GPO Access. ...
Cao, Binbin; Yan, Huifang; Guo, Mangmang; Xie, Han; Wu, Ye; Gu, Qiang; Xiao, Jiangxi; Shang, Jing; Yang, Yanling; Xiong, Hui; Niu, Zhengping; Wu, Xiru; Jiang, Yuwu; Wang, Jingmin
2016-01-01
Objective Megalencephalic leukoencephalopathy with subcortical cysts (MLC, OMIM 604004) is a rare neurological deterioration disease. We aimed to clarify clinical and genetic features of Chinese MLC patients. Methods Clinical information and peripheral venous blood of 20 patients and their families were collected, Sanger-sequencing and Multiple Ligation-dependent Probe Amplification were performed to make genetic analysis. Splicing-site mutation was confirmed with RT-PCR. UPD was detected by haplotype analysis. Follow-up study was performed through telephone for 27 patients. Results Out of 20 patients, macrocephaly, classic MRI features, motor development delay and cognitive impairment were detected in 20(100%), 20(100%), 17(85%) and 4(20%) patients, respectively. 20(100%) were clinically diagnosed with MLC. 19(95%) were genetically diagnosed with 10 novel mutations in MLC1, MLC1 and GlialCAM mutations were identified in 15 and 4 patients, respectively. Deletion mutation from exon4 to exon9 and a homozygous point mutation due to maternal UPD of chromosome22 in MLC1 were found firstly. c.598-2A>C in MLC1 leads to the skip of exon8. c.772-1G>C in MLC1 accounting for 15.5%(9/58) alleles in Chinese patients might be a founder or a hot-spot mutation. Out of 27 patients in the follow-up study, head circumference was ranged from 56cm to 61cm in patients older than 5yeas old, with a median of 57cm. Motor development delay and cognitive impairment were detected in 22(81.5%) and 5(18.5%) patients, respectively. Motor and cognitive deterioration was found in 5 (18.5%) and 2 patients (7.4%), respectively. Improvements and MRI recovery were first found in Chinese patients. Rate of seizures (45.5%), transient motor retrogress (45.5%) and unconsciousness (13.6%) after head trauma was much higher than that after fever (18.2%, 9.1%, 0%, respectively). Significance It’s a clinical and genetic analysis and a follow-up study for largest sample of Chinese MLC patients, identifying 10 novel mutations, expanding mutation spectrums and discovering clinical features of Chinese MLC patients. PMID:27322623
Real-time tracking of respiratory-induced tumor motion by dose-rate regulation
NASA Astrophysics Data System (ADS)
Han-Oh, Yeonju Sarah
We have developed a novel real-time tumor-tracking technology, called Dose-Rate-Regulated Tracking (DRRT), to compensate for tumor motion caused by breathing. Unlike other previously proposed tumor-tracking methods, this new method uses a preprogrammed dynamic multileaf collimator (MLC) sequence in combination with real-time dose-rate control. This new scheme circumvents the technical challenge in MLC-based tumor tracking, that is to control the MLC motion in real time, based on real-time detected tumor motion. The preprogrammed MLC sequence describes the movement of the tumor, as a function of breathing phase, amplitude, or tidal volume. The irregularity of tumor motion during treatment is handled by real-time regulation of the dose rate, which effectively speeds up or slows down the delivery of radiation as needed. This method is based on the fact that all of the parameters in dynamic radiation delivery, including MLC motion, are enslaved to the cumulative dose, which, in turn, can be accelerated or decelerated by varying the dose rate. Because commercially available MLC systems do not allow the MLC delivery sequence to be modified in real time based on the patient's breathing signal, previously proposed tumor-tracking techniques using a MLC cannot be readily implemented in the clinic today. By using a preprogrammed MLC sequence to handle the required motion, the task for real-time control is greatly simplified. We have developed and tested the pre- programmed MLC sequence and the dose-rate regulation algorithm using lung-cancer patients breathing signals. It has been shown that DRRT can track the tumor with an accuracy of less than 2 mm for a latency of the DRRT system of less than 0.35 s. We also have evaluated the usefulness of guided breathing for DRRT. Since DRRT by its very nature can compensate for breathing-period changes, guided breathing was shown to be unnecessary for real-time tracking when using DRRT. Finally, DRRT uses the existing dose-rate control system that is provided for current linear accelerators. Therefore, DRRT can be achieved with minimal modification of existing technology, and this can shorten substantially the time necessary to establish DRRT in clinical practice.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Colvill, Emma; Northern Sydney Cancer Centre, Royal North Shore Hospital, St. Leonards, NSW; Booth, Jeremy T.
2015-08-01
Purpose: To test the hypothesis that multileaf collimator (MLC) tracking improves the consistency between the planned and delivered dose compared with the dose without MLC tracking, in the setting of a prostate cancer volumetric modulated arc therapy trial. Methods and Materials: Multileaf collimator tracking was implemented for 15 patients in a prostate cancer radiation therapy trial; in total, 513 treatment fractions were delivered. During each treatment fraction, the prostate trajectory and treatment MLC positions were collected. These data were used as input for dose reconstruction (multiple isocenter shift method) to calculate the treated dose (with MLC tracking) and the dose thatmore » would have been delivered had MLC tracking not been applied (without MLC tracking). The percentage difference from planned for target and normal tissue dose-volume points were calculated. The hypothesis was tested for each dose-volume value via analysis of variance using the F test. Results: Of the 513 fractions delivered, 475 (93%) were suitable for analysis. The mean difference and standard deviation between the planned and treated MLC tracking doses and the planned and without-MLC tracking doses for all 475 fractions were, respectively, PTV D{sub 99%} −0.8% ± 1.1% versus −2.1% ± 2.7%; CTV D{sub 99%} −0.6% ± 0.8% versus −0.6% ± 1.1%; rectum V{sub 65%} 1.6% ± 7.9% versus −1.2% ± 18%; and bladder V{sub 65%} 0.5% ± 4.4% versus −0.0% ± 9.2% (P<.001 for all dose-volume results). Conclusion: This study shows that MLC tracking improves the consistency between the planned and delivered doses compared with the modeled doses without MLC tracking. The implications of this finding are potentially improved patient outcomes, as well as more reliable dose-volume data for radiobiological parameter determination.« less
1994-01-01
Mechanisms of Ca2+ sensitization of both myosin light chain (MLC) phosphorylation and force development by protein kinase C (PKC) were studied in permeabilized tonic smooth muscle obtained from the rabbit femoral artery. For comparison, the Ca2+ sensitizing effect of guanosine 5'-O-(gamma-thiotriphosphate) (GTP gamma S) was examined, which had been previously shown to inhibit MLC phosphatase in phasic vascular smooth muscle. We now report that PKC activators (phorbol esters, short chain synthetic diacylglycerols and a diacylglycerol kinase inhibitor) and GTP gamma S significantly increase both MLC phosphorylation and force development at constant [Ca2+]. Major phosphorylation site occurring in the presence of phorbol-12,13- dibutyrate (PDBu) or GTP gamma S at constant [Ca2+] is the same serine residue (Ser-19) as that phosphorylated by MLC kinase in response to increased Ca2+ concentrations. In an ATP- and Ca(2+)-free solution containing 1-(5-chloronaphthalene-1-sulfonyl)-1H-hexahydro-1,4- diazepine (ML-9), to avoid the kinase activity, both PDBu and GTP gamma S significantly decreased the rate of MLC dephosphorylation to half its control value. However, PDBu inhibited the relaxation rate more than did GTP gamma S. In the presence of microcystin-LR to inhibit the phosphatase activity, neither PDBu nor GTP gamma S affected MLC phosphorylation and force development. These results indicate that PKC, like activation of GTP binding protein, increases Ca2+ sensitivity of both MLC phosphorylation and force production through inhibition of MLC phosphatase. PMID:7807049
The Mathematics and Computer Science Learning Center (MLC).
ERIC Educational Resources Information Center
Abraham, Solomon T.
The Mathematics and Computer Science Learning Center (MLC) was established in the Department of Mathematics at North Carolina Central University during the fall semester of the 1982-83 academic year. The initial operations of the MLC were supported by grants to the University from the Burroughs-Wellcome Company and the Kenan Charitable Trust Fund.…
Federal Register 2010, 2011, 2012, 2013, 2014
2011-01-25
... INTERNATIONAL TRADE COMMISSION [Investigation No. 337-TA-683] In the Matter of Certain MLC Flash Memory Devices and Products Containing Same; Notice of Commission Determination Not To Review an Initial... the United States after importation of certain MLC flash memory devices and products containing same...
DOE Office of Scientific and Technical Information (OSTI.GOV)
Nguyen, A; Ironwood CRC, Phoenix, AZ; Rajaguru, P
2014-06-15
Purpose: To establish a set of tests based on the iCOM software that can be used to commission and perform periodic QA of VMAT delivery on the Elekta Synergy-S, commonly known as the Beam Modulator (BM). Methods: iCOM is used to create and deliver customized treatment fields to characterize the system in terms of 1) MLC positioning accuracy under static and dynamic delivery with full gantry rotation, 2) MLC positioning with known errors, 3) Maximum dose rate, 4) Maximum MLC speed, 5) Maximum gantry speed, 6) Synchronization: gantry speed versus dose rate, and 7) Synchronization: MLC speed versus dose rate.more » The resulting images were captured on the iView GT and exported in DICOM format to Dosimetry Check™ system for visual and quantitative analysis. For the initial commissioning phase, the system tests described should be supplemented with extensive patient QAs covering all clinically relevant treatment sites. Results: The system performance test suite showed that on our Synergy-S, MLC positioning was accurate under both static and dynamic deliveries. Intentional errors of 1 mm were also easily identified on both static and dynamic picket fence tests. Maximum dose rate was verified with stop watch to be consistently between 475-480 MU/min. Maximum gantry speed and MLC speed were 5.5 degree/s and 2.5 cm/s respectively. After accounting for beam flatness, both synchronization tests, gantry versus dose rate and MLC speed versus dose rate, were successful as the fields were uniform across the strips and there were no obvious cold/hot spots. Conclusion: VMAT commissioning and quality assurance should include machine characterization tests in addition to patient QAs. Elekta iCOM is a valuable tool for the design of customized VMAT field with specific MU, MLC leaf positions, dose rate, and indirect control of MLC and gantry speed at each of its control points.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Ganesh, T; Sarkar, B; Munshi, A
2016-06-15
Purpose: Objective of this study was to evaluate the impact of using flattening filter free (FFF) beam with 0.5 cm multileaf collimator (MLC) leaves over conventional flattened beam with 1 cm leaf width MLC on the treatment plan quality in cranio-spinal irradiation (CSI). Methods: For five medulloblastoma cases (3 males and 2 females), who were previously treated by volumetric modulated arc therapy (VMAT) technique using conventional flattened beam shaped by 1 cm width MLC leaves, four test plans were generated and compared against the delivered plan. These retrospective plans consisted of four different combinations of flattened and FFF beams frommore » Elekta’s Agility treatment head with 0.5 cm width MLC leaves. Sparing of organs at risks (OAR) in terms of dose to 5%, 50%, 75% and 90% volumes, mean and maximum dose were evaluated. Results: All plans satisfied the planning objective of covering 95% of PTV by at least 95% of prescription dose. Marginal variation of dose spillage was observed between different VMAT plans at very low dose range (1–5 Gy). Variation in dose statistics for PTVs and OARs were within 1% or 1 Gy. Amongst the five plans, the plan with flattened beam with 1 cm MLC had the highest number of MUs, 2.13 times higher than the plan with Agility MLC with FFF beam that had the least number of MUs. No statistically significant difference (p≥0.05) was observed between the reference plan and the retrospectively generated plans in terms of PTV coverage, cold spot, hot spot and organ at risk doses. Conclusion: In the treatment of CSI cases by VMAT technique, FFF beams and/or finer width MLC did not exhibit advantage over the flattened beams or wider MLC in terms of plan quality except for reduction in MUs.« less
Wang, Tanchun; Kendig, Derek M; Smolock, Elaine M; Moreland, Robert S
2009-12-01
Smooth muscle contraction is regulated by phosphorylation of the myosin light chain (MLC) catalyzed by MLC kinase and dephosphorylation catalyzed by MLC phosphatase. Agonist stimulation of smooth muscle results in the inhibition of MLC phosphatase activity and a net increase in MLC phosphorylation and therefore force. The two pathways believed to be primarily important for inhibition of MLC phosphatase activity are protein kinase C (PKC)-catalyzed CPI-17 phosphorylation and Rho kinase (ROCK)-catalyzed myosin phosphatase-targeting subunit (MYPT1) phosphorylation. The goal of this study was to determine the roles of PKC and ROCK and their downstream effectors in regulating MLC phosphorylation levels and force during the phasic and sustained phases of carbachol-stimulated contraction in intact bladder smooth muscle. These studies were performed in the presence and absence of the PKC inhibitor bisindolylmaleimide-1 (Bis) or the ROCK inhibitor H-1152. Phosphorylation levels of Thr(38)-CPI-17 and Thr(696)/Thr(850)-MYPT1 were measured at different times during carbachol stimulation using site-specific antibodies. Thr(38)-CPI-17 phosphorylation increased concurrently with carbachol-stimulated force generation. This increase was reduced by inhibition of PKC during the entire contraction but was only reduced by ROCK inhibition during the sustained phase of contraction. MYPT1 showed high basal phosphorylation levels at both sites; however, only Thr(850) phosphorylation increased with carbachol stimulation; the increase was abolished by the inhibition of either ROCK or PKC. Our results suggest that during agonist stimulation, PKC regulates MLC phosphatase activity through phosphorylation of CPI-17. In contrast, ROCK phosphorylates both Thr(850)-MYPT1 and CPI-17, possibly through cross talk with a PKC pathway, but is only significant during the sustained phase of contraction. Last, our results demonstrate that there is a constitutively activate pool of ROCK that phosphorylates MYPT1 in the basal state, which may account for the high resting levels of MLC phosphorylation measured in rabbit bladder smooth muscle.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Haddad, K; Alopoor, H
Purpose: Recently, the multileaf collimators (MLC) have become an important part of any LINAC collimation systems because they reduce the treatment planning time and improves the conformity. Important factors that affects the MLCs collimation performance are leaves material composition and their thickness. In this study, we investigate the main dosimetric parameters of 120-leaf Millennium MLC including dose in the buildup point, physical penumbra as well as average and end leaf leakages. Effects of the leaves geometry and density on these parameters are evaluated Methods: From EGSnrc Monte Carlo code, BEAMnrc and DOSXYZnrc modules are used to evaluate the dosimetric parametersmore » of a water phantom exposed to a Varian xi for 100cm SSD. Using IAEA phasespace data just above MLC (Z=46cm) and BEAMnrc, for the modified 120-leaf Millennium MLC a new phase space data at Z=52cm is produces. The MLC is modified both in leaf thickness and material composition. EGSgui code generates 521ICRU library for tungsten alloys. DOSXYZnrc with the new phase space evaluates the dose distribution in a water phantom of 60×60×20 cm3 with voxel size of 4×4×2 mm3. Using DOSXYZnrc dose distributions for open beam and closed beam as well as the leakages definition, end leakage, average leakage and physical penumbra are evaluated. Results: A new MLC with improved dosimetric parameters is proposed. The physical penumbra for proposed MLC is 4.7mm compared to 5.16 mm for Millennium. Average leakage in our design is reduced to 1.16% compared to 1.73% for Millennium, the end leaf leakage suggested design is also reduced to 4.86% compared to 7.26% of Millennium. Conclusion: The results show that the proposed MLC with enhanced dosimetric parameters could improve the conformity of treatment planning.« less
Jin, Lihui; Eldib, Ahmed; Li, Jinsheng; Emam, Ismail; Fan, Jiajin; Wang, Lu; Ma, C-M
2014-01-06
The dosimetric advantage of modulated electron radiotherapy (MERT) has been explored by many investigators and is considered to be an advanced radiation therapy technique in the utilization of electrons. A computer-controlled electron multileaf collimator (MLC) prototype, newly designed to be added onto a Varian linac to deliver MERT, was investigated both experimentally and by Monte Carlo simulations. Four different electron energies, 6, 9, 12, and 15 MeV, were employed for this investigation. To ensure that this device was capable of delivering the electron beams properly, measurements were performed to examine the electron MLC (eMLC) leaf leakage and to determine the appropriate jaw positioning for an eMLC-shaped field in order to eliminate a secondary radiation peak that could otherwise appear outside of an intended radiation field in the case of inappropriate jaw positioning due to insufficient radiation blockage from the jaws. Phase space data were obtained by Monte Carlo (MC) simulation and recorded at the plane just above the jaws for each of the energies (6, 9, 12, and 15 MeV). As an input source, phase space data were used in MC dose calculations for various sizes of the eMLC shaped field (10 × 10 cm2, 3.4 × 3.4 cm2, and 2 × 2 cm2) with respect to a water phantom at source-to-surface distance (SSD) = 94 cm, while the jaws, eMLC leaves, and some accessories associated with the eMLC assembly as well were modeled as modifiers in the calculations. The calculated results were then compared with measurements from a water scanning system. The results showed that jaw settings with 5 mm margins beyond the field shaped by the eMLC were appropriate to eliminate the secondary radiation peak while not widening the beam penumbra; the eMLC leaf leakage measurements ranged from 0.3% to 1.8% for different energies based on in-phantom measurements, which should be quite acceptable for MERT. Comparisons between MC dose calculations and measurements showed agreement within 1%/1 mm based on percentage depth doses (PDDs) and off-axis dose profiles for a range of field sizes for each of the electron energies. Our current work has demonstrated that the eMLC and other relevant components in the linac were correctly modeled and simulated via our in-house MC codes, and the eMLC is capable of accurately delivering electron beams for various eMLC-shaped field sizes with appropriate jaw settings. In the next stage, patient-specific verification with a full MERT plan should be performed.
Ansari, Habib R; Davis, Angela M; Kaddour-Djebbar, Ismail; Abdel-Latif, Ata A
2003-06-01
The effects of the ocular hypotensive agents prostaglandin F(2alpha) (PGF(2alpha)) and its analog latanoprost on intraocular pressure (IOP) in both animals and human have been investigated extensively in the last two decades. While there is general agreement that application of these PGs to the eye alters IOP by altering the aqueous humor outflow of the eye via the uveoscleral and trabecular meshwork pathways, the mechanism of action of these agents on IOP lowering remains unclear. There is evidence which suggests that myosin light kinase (MLC kinase) may be involved in the IOP-lowering effects of these agents. Thus, the purpose of the present work was to investigate in cat iris sphincter the effects of these PGs on the MLC kinase signaling pathway, inositol phosphates production, MLC phosphorylation and contraction, in order to gain more information about the mechanism through which these agents modulate smooth muscle function and lower IOP. [(3)H]myo-inositol phosphates production was measured by ion-exchange chromatography, MLC kinase activity was measured by incorporation of (32)Pi into MLC, and changes in muscle tension were recorded isometrically. PGF(2alpha) and latanoprost induced contraction in a concentration-dependent manner with EC(50) values of 18.6 and 29.9 nM, respectively, and increased inositol phosphates production in a concentration-dependent manner. At 1 microM, PGF(2alpha) and latanoprost increased inositol phosphates formation by 125 and 102% over basal, respectively. PGF(2alpha) and latanoprost increased MLC phosphorylation in a concentration- and time-dependent manner, at 1 microM and 5 min incubation, the PGs increased the MLC response by 181 and 176% over basal, respectively. In general, PGF(2alpha) was slightly more potent in inducing the biochemical and pharmacological responses. Wortmannin, ML-7 and ML-9, selective inhibitors of MLC kinase, inhibited significantly PGF(2alpha)- and latanoprost-induced MLC phosphorylation and contraction. These results demonstrate for the first time involvement of the MLC kinase pathway in the FP receptor function of this ocular tissue. Contraction-relaxation of smooth muscle alters the shape and stiffness of smooth muscle cells and MLC kinase, through myosin phosphorylation and dephosphorylation, has been shown to be involved in cytoskeletal remodeling, cytoskeletal alterations, and IOP lowering. In light of these reports and the findings presented here we suggest that alterations in the MLC kinase signaling pathway and its derived second messengers, which leads to changes in contraction-relaxation of the smooth muscles of the anterior segment, could facilitate aqueous humor outflow and thus contribute to the IOP-lowering effects of the FP-class PGs.
NASA Astrophysics Data System (ADS)
Glitzner, M.; Crijns, S. P. M.; de Senneville, B. Denis; Lagendijk, J. J. W.; Raaymakers, B. W.
2015-03-01
For motion adaptive radiotherapy, dynamic multileaf collimator tracking can be employed to reduce treatment margins by steering the beam according to the organ motion. The Elekta Agility 160 MLC has hitherto not been evaluated for its tracking suitability. Both dosimetric performance and latency are key figures and need to be assessed generically, independent of the used motion sensor. In this paper, we propose the use of harmonic functions directly fed to the MLC to determine its latency during continuous motion. Furthermore, a control variable is extracted from a camera system and fed to the MLC. Using this setup, film dosimetry and subsequent γ statistics are performed, evaluating the response when tracking (MRI)-based physiologic motion in a closed-loop. The delay attributed to the MLC itself was shown to be a minor contributor to the overall feedback chain as compared to the impact of imaging components such as MRI sequences. Delay showed a linear phase behaviour of the MLC employed in continuously dynamic applications, which enables a general MLC-characterization. Using the exemplary feedback chain, dosimetry showed a vast increase in pass rate employing γ statistics. In this early stage, the tracking performance of the Agility using the test bench yielded promising results, making the technique eligible for translation to tracking using clinical imaging modalities.
Lee, So-Rim; Xu, Yong-Nan; Jo, Yu-Jin; Namgoong, Suk; Kim, Nam-Hyung
2015-11-01
Oocyte meiosis involves a unique asymmetric division involving spindle movement from the central cytoplasm to the cortex, followed by polar body extrusion. ROCK is a Rho-GTPase effector involved in various cellular functions in somatic cells as well as oocyte meiosis. ROCK was previously shown to promote actin organization by phosphorylating several downstream targets, including LIM domain kinase (LIMK), phosphorylated cofilin (p-cofilin), and myosin light chain (MLC). In this study, we investigated the roles of ROCK and MLC during bovine oocyte meiosis. We found that ROCK was localized around the nucleus at the oocyte's germinal-vesicle (GV) stage, but spreads to the rest of the cytoplasm in later developmental stages. On the other hand, phosphorylated MLC (p-MLC) localized at the cortex, and its abundance decreased by the metaphase-II stage. Disrupting ROCK activity, via RNAi or the chemical inhibitor Y-27632, blocked both cell cycle progression and polar body extrusion. ROCK inhibition also resulted in decreased cortical actin, p-cofilin, and p-MLC levels. Similar to the phenotype associated with inhibition of ROCK activity, inhibition of MLC kinase by the chemical inhibitor ML-7 caused defects in polar body extrusion. Collectively, our results suggest that the ROCK/MLC/actomyosin as well as ROCK/LIMK/cofilin pathways regulate meiotic spindle migration and cytokinesis during bovine oocyte maturation. © 2015 Wiley Periodicals, Inc.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Owen, R. K.
2007-04-04
A perl module designed to read and parse the voluminous set of event or accounting log files produced by a Portable Batch System (PBS) server. This module can filter on date-time and/or record type. The data can be returned in a variety of formats.
Expansion of the roadway reference log : KYSPR-99-201.
DOT National Transportation Integrated Search
2000-05-01
The objectives of this study were to: 1) expand the current route log to include milepoints for all intersections on state maintained roads and 2) recommend a procedure for establishing milepoints and maintaining the file with up-to-date information....
78 FR 52524 - Sunoco Pipeline LP; Notice of Petition for Declaratory Order
Federal Register 2010, 2011, 2012, 2013, 2014
2013-08-23
... link to log on and submit the intervention or protests. Persons unable to file electronically should... described in their petition. Any person desiring to intervene or to protest in this proceedings must file in... service, persons with Internet access who will eFile a document and/or be listed as a contact for an...
78 FR 62349 - Sunoco Pipeline L.P.; Notice of Petition for Declaratory Order
Federal Register 2010, 2011, 2012, 2013, 2014
2013-10-18
... to log on and submit the intervention or protests. Persons unable to file electronically should... petition. Any person desiring to intervene or to protest in this proceeding must file in accordance with..., persons with Internet access who will eFile a document and/or be listed as a contact for an intervenor...
17 CFR 249.446 - Form ID, uniform application for access codes to file on EDGAR.
Code of Federal Regulations, 2010 CFR
2010-04-01
... log on to the EDGAR system, submit filings, and change its CCC. (d) Password Modification Authorization Code (PMAC)—allows a filer, filing agent or training agent to change its Password. [69 FR 22710... Sections Affected, which appears in the Finding Aids section of the printed volume and on GPO Access. ...
Federal Register 2010, 2011, 2012, 2013, 2014
2013-12-20
... through DEMD's in-house databases; Well log interpretation, including correlation of formation tops.... Files must have descriptive file names to help DEMD quickly locate specific components of the proposal...
Gannon, Joan; Doran, Philip; Kirwan, Anne; Ohlendieck, Kay
2009-11-01
The age-dependent decline in skeletal muscle mass and function is believed to be due to a multi-factorial pathology and represents a major factor that blocks healthy aging by increasing physical disability, frailty and loss of independence in the elderly. This study has focused on the comparative proteomic analysis of contractile elements and revealed that the most striking age-related changes seem to occur in the protein family representing myosin light chains (MLCs). Comparative screening of total muscle extracts suggests a fast-to-slow transition in the aged MLC population. The mass spectrometric analysis of the myofibril-enriched fraction identified the MLC2 isoform of the slow-type MLC as the contractile protein with the most drastically changed expression during aging. Immunoblotting confirmed an increased abundance of slow MLC2, concomitant with a switch in fast versus slow myosin heavy chains. Staining of two-dimensional gels of crude extracts with the phospho-specific fluorescent dye ProQ-Diamond identified the increased MLC2 spot as a muscle protein with a drastically enhanced phosphorylation level in aged fibres. Comparative immunofluorescence microscopy, using antibodies to fast and slow myosin isoforms, confirmed a fast-to-slow transformation process during muscle aging. Interestingly, the dramatic increase in slow MLC2 expression was restricted to individual senescent fibres. These findings agree with the idea that aged skeletal muscles undergo a shift to more aerobic-oxidative metabolism in a slower-twitching fibre population and suggest the slow MLC2 isoform as a potential biomarker for fibre type shifting in sarcopenia of old age.
SU-F-T-313: Clinical Results of a New Customer Acceptance Test for Elekta VMAT
DOE Office of Scientific and Technical Information (OSTI.GOV)
Rusk, B; Fontenot, J
Purpose: To report the results of a customer acceptance test (CAT) for VMAT treatments for two matched Elekta linear accelerators. Methods: The CAT tests were performed on two clinically matched Elekta linear accelerators equipped with a 160-leaf MLC. Functional tests included performance checks of the control system during dynamic movements of the diaphragms, MLC, and gantry. Dosimetric tests included MLC picket fence tests at static and variable dose rates and a diaphragm alignment test, all performed using the on-board EPID. Additionally, beam symmetry during arc delivery was measured at the four cardinal angles for high and low dose rate modesmore » using a 2D detector array. Results of the dosimetric tests were analyzed using the VMAT CAT analysis tool. Results: Linear accelerator 1 (LN1) met all stated CAT tolerances. Linear accelerator 2 (LN2) passed the geometric, beam symmetry, and MLC position error tests but failed the relative dose average test for the diaphragm abutment and all three picket fence fields. Though peak doses in the abutment regions were consistent, the average dose was below the stated tolerance corresponding to a leaf junction that was too narrow. Despite this, no significant differences in patient specific VMAT quality assurance measured were observed between the accelerators and both passed monthly MLC quality assurance performed with the Hancock test. Conclusion: Results from the CAT showed LN2 with relative dose averages in the abutment regions of the diaphragm and MLC tests outside the tolerances resulting from differences in leaf gap distances. Tolerances of the dose average tests from the CAT may be small enough to detect MLC errors which do not significantly affect patient QA or the routine MLC tests.« less
Building analytical platform with Big Data solutions for log files of PanDA infrastructure
NASA Astrophysics Data System (ADS)
Alekseev, A. A.; Barreiro Megino, F. G.; Klimentov, A. A.; Korchuganova, T. A.; Maendo, T.; Padolski, S. V.
2018-05-01
The paper describes the implementation of a high-performance system for the processing and analysis of log files for the PanDA infrastructure of the ATLAS experiment at the Large Hadron Collider (LHC), responsible for the workload management of order of 2M daily jobs across the Worldwide LHC Computing Grid. The solution is based on the ELK technology stack, which includes several components: Filebeat, Logstash, ElasticSearch (ES), and Kibana. Filebeat is used to collect data from logs. Logstash processes data and export to Elasticsearch. ES are responsible for centralized data storage. Accumulated data in ES can be viewed using a special software Kibana. These components were integrated with the PanDA infrastructure and replaced previous log processing systems for increased scalability and usability. The authors will describe all the components and their configuration tuning for the current tasks, the scale of the actual system and give several real-life examples of how this centralized log processing and storage service is used to showcase the advantages for daily operations.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Eldib, A; Al-Azhar University Cairo; Jin, L
2014-06-01
Purpose: Modulated electron radiotherapy (MERT) has the potential to achieve better treatment outcome for shallow tumors such as those of breast and scalp. In a separate study with scalp lesions, MERT was compared to volumetric modulated arc therapy. Our results showed a reduction in the dose reaching the brain with MERT. However dose calculation accuracy and delivery efficiency challenges remain. Thus in the current study we proceed to add more cases to demonstrate MERT beneficial outcome and its delivery accuracy using an electron specific multileaf collimator (eMLC). Methods: We have used the MCBEAM code for treatment head simulation and formore » generating phase space files to be used as radiation source input for our Monte Carlo based treatment planning system (MC TPS). MCPLAN code is used for calculation of patient specific dose deposition coefficient and for final MERT plan dose calculation. An in-house developed optimization code is used for the optimization process. MERT plans were generated for real patients and head and neck phantom. Film was used for dosimetric verification. The film was cut following the contour of the curved phantom surface and then sealed with black masking tape. In the measurement, the sealed film packet was sandwiched between two adjacent slabs of the head and neck phantom. The measured 2D dose distribution was then compared with calculations. Results: The eMLC allows effective treatment of scalps with multi-lesions spreading around the patient head, which was usually difficult to plan or very time consuming with conventional applicators. MERT continues to show better reduction in the brain dose. The dosimetric measurements showed slight discrepancy, which was attributed to the film setup. Conclusion: MERT can improve treatment plan quality for patients with scalp cancers. Our in-house MC TPS is capable of performing treatment planning and accurate dose calculation for MERT using the eMLC.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Hussain, A
Purpose: Novel linac machines, TrueBeam (TB) and Elekta Versa have updated head designing and software control system, include flattening-filter-free (FFF) photon and electron beams. Later on FFF beams were also introduced on C-Series machines. In this work FFF beams for same energy 6MV but from different machine versions were studied with reference to beam data parameters. Methods: The 6MV-FFF percent depth doses, profile symmetry and flatness, dose rate tables, and multi-leaf collimator (MLC) transmission factors were measured during commissioning process of both C-series and Truebeam machines. The scanning and dosimetric data for 6MV-FFF beam from Truebeam and C-Series linacs wasmore » compared. A correlation of 6MV-FFF beam from Elekta Versa with that of Varian linacs was also found. Results: The scanning files were plotted for both qualitative and quantitative analysis. The dosimetric leaf gap (DLG) for C-Series 6MV-FFF beam is 1.1 mm. Published values for Truebeam dosimetric leaf gap is 1.16 mm. 6MV MLC transmission factor varies between 1.3 % and 1.4 % in two separate measurements and measured DLG values vary between 1.32 mm and 1.33 mm on C-Series machine. MLC transmission factor from C-Series machine varies between 1.5 % and 1.6 %. Some of the measured data values from C-Series FFF beam are compared with Truebeam representative data. 6MV-FFF beam parameter values like dmax, OP factors, beam symmetry and flatness and additional parameters for C-Series and Truebeam liancs will be presented and compared in graphical form and tabular data form if selected. Conclusion: The 6MV flattening filter (FF) beam data from C-Series & Truebeam and 6MV-FFF beam data from Truebeam has already presented. This particular analysis to compare 6MV-FFF beam from C-Series and Truebeam provides opportunity to better elaborate FFF mode on novel machines. It was found that C-Series and Truebeam 6MV-FFF dosimetric and beam data was quite similar.« less
Wang, Lilie; Ding, George X
2018-06-12
Therapeutic radiation to cancer patients is accompanied by unintended radiation to organs outside the treatment field. It is known that the model-based dose algorithm has limitation in calculating the out-of-field doses. This study evaluated the out-of-field dose calculated by the Varian Eclipse treatment planning system (v.11 with AAA algorithm) in realistic treatment plans with the goal of estimating the uncertainties of calculated organ doses. Photon beam phase-space files for TrueBeam linear accelerator were provided by Varian. These were used as incident sources in EGSnrc Monte Carlo simulations of radiation transport through the downstream jaws and MLC. Dynamic movements of the MLC leaves were fully modeled based on treatment plans using IMRT or VMAT techniques. The Monte Carlo calculated out-of-field doses were then compared with those calculated by Eclipse. The dose comparisons were performed for different beam energies and treatment sites, including head-and-neck, lung, and pelvis. For 6 MV (FF/FFF), 10 MV (FF/FFF), and 15 MV (FF) beams, Eclipse underestimated out-of-field local doses by 30%-50% compared with Monte Carlo calculations when the local dose was <1% of prescribed dose. The accuracy of out-of-field dose calculations using Eclipse is improved when collimator jaws were set at the smallest possible aperture for MLC openings. The Eclipse system consistently underestimates out-of-field dose by a factor of 2 for all beam energies studied at the local dose level of less than 1% of prescribed dose. These findings are useful in providing information on the uncertainties of out-of-field organ doses calculated by Eclipse treatment planning system. © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.
ERIC Educational Resources Information Center
Xu, Cora Lingling
2017-01-01
Drawing on in-depth interview data from 31 mainland Chinese (MLC) students in a Hong Kong university, this article conceptualises MLC and Hong Kong higher education as two dissonant but interrelated subfields of the Chinese higher education field. The article argues that these MLC students' habitus, one that possesses rich economic, social and…
SU-E-J-67: Evaluation of Adaptive MLC Morphing for Online Correction of Prostate Cancer Radiotherapy
DOE Office of Scientific and Technical Information (OSTI.GOV)
Sandhu, R; Qin, A; Yan, D
Purpose: Online adaptive MLC morphing is desirable over translational couch shifts to accommodate target position as well as anatomic changes. A reliable method of adaptive MLC segment to target during prostate cancer IMRT treatment is proposed and evaluated by comparison with daily online-image guidance (IGRT) correction and online-IMRT planning. Methods: The MLC adaptive algorithm involves following steps; move the MLC segments according to target translational shifts, and then morph the segment shape to maintain the spatial relationship between the planning-target contour and MLC segment. Efficacy of this method was evaluated retrospectively using daily-CBCT images on seven prostate patients treated withmore » seven-beam IMRT treatment to deliver 64Gy in 20 fractions. Daily modification was simulated with three approaches; daily-IGRT correction based on implanted radio-markers, adaptive MLC morphing, and online-IMRT planning, with no-residual variation. The selected dosimetric endpoints and nEUD (normalized equivalent uniform dose to online-IMRT planning) of each organ of interest were determined for evaluation and comparison. Results: For target(prostate), bladder and rectal-wall, the mean±sd of nEUD were 97.6%+3.2%, 103.9%±4.9% and 97.4%±1.1% for daily-IGRT correction; and 100.2%+0.2%, 108.9%±5.1% and 99.8%±1.2% for adaptive MLC morphing, respectively. For daily-IGRT correction, adaptive MLC morphing and online-IMRT planning, target D99 was <95% of the prescription dose in 30%, 0% and 0% of 140 fractions, respectively. For the rectal-wall, D5 exceeded 105% of the planned-D5 in 2.8%, 11.4% and 0% of 140 fractions, respectively. For the bladder, Dmax exceeded 105% of the planned-D5 in 2.8%, 5.6% and 0% of 140 fractions, respectively. D30 of bladder and rectal-wall were well within the planned-D30 for all three approaches. Conclusion: The proposed method of adaptive MLC morphing can be beneficial for the prostate patient population with large deformation and rotation. It is superior to the daily-IGRT correction, and comparable to the online-IMRT planning for dose to the target and rectal-wall.« less
20 CFR 658.414 - Referral of non-JS-related complaints.
Code of Federal Regulations, 2011 CFR
2011-04-01
... applicable, were referred on the complaint log specified in § 658.410(c)(1). The JS official shall also prepare and keep the file specified in § 658.410(c)(3) for the complaints filed pursuant to paragraph (a...
Booth, Jeremy T; Caillet, Vincent; Hardcastle, Nicholas; O'Brien, Ricky; Szymura, Kathryn; Crasta, Charlene; Harris, Benjamin; Haddad, Carol; Eade, Thomas; Keall, Paul J
2016-10-01
Real time adaptive radiotherapy that enables smaller irradiated volumes may reduce pulmonary toxicity. We report on the first patient treatment of electromagnetic-guided real time adaptive radiotherapy delivered with MLC tracking for lung stereotactic ablative body radiotherapy. A clinical trial was developed to investigate the safety and feasibility of MLC tracking in lung. The first patient was an 80-year old man with a single left lower lobe lung metastasis to be treated with SABR to 48Gy in 4 fractions. In-house software was integrated with a standard linear accelerator to adapt the treatment beam shape and position based on electromagnetic transponders implanted in the lung. MLC tracking plans were compared against standard ITV-based treatment planning. MLC tracking plan delivery was reconstructed in the patient to confirm safe delivery. Real time adaptive radiotherapy delivered with MLC tracking compared to standard ITV-based planning reduced the PTV by 41% (18.7-11cm 3 ) and the mean lung dose by 30% (202-140cGy), V20 by 35% (2.6-1.5%) and V5 by 9% (8.9-8%). An emerging technology, MLC tracking, has been translated into the clinic and used to treat lung SABR patients for the first time. This milestone represents an important first step for clinical real-time adaptive radiotherapy that could reduce pulmonary toxicity in lung radiotherapy. Copyright © 2016 Elsevier Ireland Ltd. All rights reserved.
Federal Register 2010, 2011, 2012, 2013, 2014
2013-08-14
... intervene or to protest should file with the Federal Energy Regulatory Commission, 888 First Street NE... . To facilitate electronic service, persons with Internet access who will eFile a document and/or be...Registration link. Select the eFiling link to log on and submit the intervention or protests. Persons unable to...
78 FR 63977 - Enable Bakken Crude Services, LLC; Notice of Request For Waiver
Federal Register 2010, 2011, 2012, 2013, 2014
2013-10-25
... person desiring to intervene or to protest in this proceedings must file in accordance with Rules 211 and... Internet access who will eFile a document and/or be listed as a contact for an intervenor must create and validate an eRegistration account using the eRegistration link. Select the eFiling link to log on and...
17 CFR 249.446 - Form ID, uniform application for access codes to file on EDGAR.
Code of Federal Regulations, 2011 CFR
2011-04-01
... log on to the EDGAR system, submit filings, and change its CCC. (d) Password Modification Authorization Code (PMAC)—allows a filer, filing agent or training agent to change its Password. [69 FR 22710... Sections Affected, which appears in the Finding Aids section of the printed volume and at at www.fdsys.gov. ...
MAIL LOG, program theory, volume 2
NASA Technical Reports Server (NTRS)
Harris, D. K.
1979-01-01
Information relevant to the MAIL LOG program theory is documented. The L-files for mail correspondence, design information release/report, and the drawing/engineering order are given. In addition, sources for miscellaneous external routines and special support routines are documented along with a glossary of terms.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Shiinoki, T; Hanazawa, H; Shibuya, K
Purpose: The respirato ry gating system combined the TrueBeam and a new real-time tumor-tracking radiotherapy system (RTRT) was installed. The RTRT system consists of two x-ray tubes and color image intensifiers. Using fluoroscopic images, the fiducial marker which was implanted near the tumor was tracked and was used as the internal surrogate for respiratory gating. The purposes of this study was to develop the verification technique of the respiratory gating with the new RTRT using cine electronic portal image device images (EPIDs) of TrueBeam and log files of the RTRT. Methods: A patient who underwent respiratory gated SBRT of themore » lung using the RTRT were enrolled in this study. For a patient, the log files of three-dimensional coordinate of fiducial marker used as an internal surrogate were acquired using the RTRT. Simultaneously, the cine EPIDs were acquired during respiratory gated radiotherapy. The data acquisition was performed for one field at five sessions during the course of SBRT. The residual motion errors were calculated using the log files (E{sub log}). The fiducial marker used as an internal surrogate into the cine EPIDs was automatically extracted by in-house software based on the template-matching algorithm. The differences between the the marker positions of cine EPIDs and digitally reconstructed radiograph were calculated (E{sub EPID}). Results: Marker detection on EPID using in-house software was influenced by low image contrast. For one field during the course of SBRT, the respiratory gating using the RTRT showed the mean ± S.D. of 95{sup th} percentile E{sub EPID} were 1.3 ± 0.3 mm,1.1 ± 0.5 mm,and those of E{sub log} were 1.5 ± 0.2 mm, 1.1 ± 0.2 mm in LR and SI directions, respectively. Conclusion: We have developed the verification method of respiratory gating combined TrueBeam and new real-time tumor-tracking radiotherapy system using EPIDs and log files.« less
Blackcomb: Hardware-Software Co-design for Non-Volatile Memory in Exascale Systems
DOE Office of Scientific and Technical Information (OSTI.GOV)
Schreiber, Robert
Summary of technical results of Blackcomb Memory Devices We explored various different memory technologies (STTRAM, PCRAM, FeRAM, and ReRAM). The progress can be classified into three categories, below. Modeling and Tool Releases Various modeling tools have been developed over the last decade to help in the design of SRAM or DRAM-based memory hierarchies. To explore new design opportunities that NVM technologies can bring to the designers, we have developed similar high-level models for NVM, including PCRAMsim [Dong 2009], NVSim [Dong 2012], and NVMain [Poremba 2012]. NVSim is a circuit-level model for NVM performance, energy, and area estimation, which supports variousmore » NVM technologies, including STT-RAM, PCRAM, ReRAM, and legacy NAND Flash. NVSim is successfully validated against industrial NVM prototypes, and it is expected to help boost architecture-level NVM-related studies. On the other side, NVMain is a cycle accurate main memory simulator designed to simulate emerging nonvolatile memories at the architectural level. We have released these models as open source tools and provided contiguous support to them. We also proposed PS3-RAM, which is a fast, portable and scalable statistical STT-RAM reliability analysis model [Wen 2012]. Design Space Exploration and Optimization With the support of these models, we explore different device/circuit optimization techniques. For example, in [Niu 2012a] we studied the power reduction technique for the application of ECC scheme in ReRAM designs and proposed to use ECC code to relax the BER (Bit Error Rate) requirement of a single memory to improve the write energy consumption and latency for both 1T1R and cross-point ReRAM designs. In [Xu 2011], we proposed a methodology to design STT-RAM for different optimization goals such as read performance, write performance and write energy by leveraging the trade-off between write current and write time of MTJ. We also studied the tradeoffs in building a reliable crosspoint ReRAM array [Niu 2012b]. We have conducted an in depth analysis of the circuit and system level design implications of multi-layer cross-point Resistive RAM (MLCReRAM) from performance, power and reliability perspectives [Xu 2013]. The objective of this study is to understand the design trade-offs of this technology with respect to the MLC Phase Change Memory (MLCPCM).Our MLC ReRAM design at the circuit and system levels indicates that different resistance allocation schemes, programming strategies, peripheral designs, and material selections profoundly affect the area, latency, power, and reliability of MLC ReRAM. Based on this analysis, we conduct two case studies: first we compare MLC ReRAM design against MLC phase-change memory (PCM) and multi-layer cross-point ReRAM design, and point out why multi-level ReRAM is appealing; second we further explore the design space for MLC ReRAM. Architecture and Application We explored hybrid checkpointing using phase-change memory for future exascale systems [Dong 2011] and showed that the use of nonvolatile memory for local checkpointing significantly increases the number of faults covered by local checkpoints and reduces the probability of a global failure in the middle of a global checkpoint to less than 1%. We also proposed a technique called i2WAP to mitigate the write variations in NVM-based last-level cache for the improvement of the NVM lifetime [Wang 2013]. Our wear leveling technique attempts to work around the limitations of write endurance by arranging data access so that write operations can be distributed evenly across all the storage cells. During our intensive research on fault-tolerant NVM design, we found that ECC cannot effectively tolerate hard errors from limited write endurance and process imperfection. Therefore, we devised a novel Point and Discard (PAD) architecture in in [ 2012] as a hard-error-tolerant architecture for ReRAM-based Last Level Caches. PAD improves the lifetime of ReRAM caches by 1.6X-440X under different process variations without performance overhead in the system's early life. We have investigated the applicability of NVM for persistent memory design [Zhao 2013]. New byte addressable NVM enables fast persistent memory that allows in-memory persistent data objects to be updated with much higher throughput. Despite the significant improvement, the performance of these designs is only 50% of the native system with no persistence support, due to the logging or copy-on-write mechanisms used to update the persistent memory. A challenge in this approach is therefore how to efficiently enable atomic, consistent, and durable updates to ensure data persistence that survives application and/or system failures. We have designed a persistent memory system, called Klin, that can provide performance as close as that of the native system. The Klin design adopts a non-volatile cache and a non-volatile main memory for constructing a multi-versioned durable memory system, enabling atomic updates without logging or copy-on-write. Our evaluation shows that the proposed Kiln mechanism can achieve up to 2X of performance improvement to NVRAM-based persistent memory employing write-ahead logging. In addition, our design has numerous practical advantages: a simple and intuitive abstract interface, microarchitecture-level optimizations, fast recovery from failures, and no redundant writes to slow non-volatile storage media. The work was published in MICRO 2013 and received Best Paper Honorable Mentioned Award.« less
NASA Astrophysics Data System (ADS)
Hernandez, Victor; Vera-Sánchez, Juan Antonio; Vieillevigne, Laure; Saez, Jordi
2017-08-01
Adequate modelling of the multi-leaf collimator (MLC) by treatment planning systems (TPS) is essential for accurate dose calculations in intensity-modulated radiation-therapy. For this reason modern TPSs incorporate MLC characteristics such as the leaf end curvature, MLC transmission and the tongue-and-groove. However, the modelling of the tongue-and-groove is often neglected during TPS commissioning and it is not known how accurate it is. This study evaluates the dosimetric consequences of the tongue-and-groove effect for two different MLC models using both film dosimetry and ionisation chambers. A set of comprehensive tests are presented that evaluate the ability of TPSs to accurately model this effect in (a) static fields, (b) sliding window beams and (c) VMAT arcs. The tests proposed are useful for the commissioning of TPSs and for the validation of major upgrades. With the ECLIPSE TPS, relevant differences were found between calculations and measurements for beams with dynamic MLCs in the presence of the TG effect, especially for the High Definition MLC, small gap sizes and the 1 mm calculation grid. For this combination, dose differences as high as 10% and 7% were obtained for dynamic MLC gaps of 5 mm and 10 mm, respectively. These differences indicate inadequate modelling of the tongue-and-groove effect, which might not be identified without the proposed tests. In particular, the TPS tended to underestimate the calculated dose, which may require tuning of other configuration parameters in the TPS (such as the dosimetric leaf gap) in order to maximise the agreement between calculations and measurements in clinical plans. In conclusion, a need for better modelling of the MLC by TPSs is demonstrated, one of the relevant aspects being the tongue-and-groove effect. This would improve the accuracy of TPS calculations, especially for plans using small MLC gaps, such as plans with small target volumes or high complexities. Improved modelling of the MLC would also reduce the need for tuning parameters in the TPS, facilitating a more comprehensive configuration and commissioning of TPSs.
Aude, Y Wady; Agatston, Arthur S; Lopez-Jimenez, Francisco; Lieberman, Eric H; Marie Almon; Hansen, Melinda; Rojas, Gerardo; Lamas, Gervasio A; Hennekens, Charles H
2004-10-25
In the United States, obesity is a major clinical and public health problem causing diabetes, dyslipidemia, and hypertension, as well as increasing cardiovascular and total mortality. Dietary restrictions of calories and saturated fat are beneficial. However, it remains unclear whether replacement of saturated fat with carbohydrates (as in the US National Cholesterol Education Program [NCEP] diet) or protein and monounsaturated fat (as in our isocaloric modified low-carbohydrate [MLC] diet, which is lower in total carbohydrates but higher in protein, monounsaturated fat, and complex carbohydrates) is optimal. We randomized 60 participants (29 women and 31 men) to the NCEP or the MLC diet and evaluated them every 2 weeks for 12 weeks. They were aged 28 to 71 years (mean age, 44 years in the NCEP and 46 years in the MLC group). A total of 36% of participants from the NCEP group and 35% from the MLC group had a body mass index (calculated as weight in kilograms divided by the square of height in meters) greater than 27. The primary end point was weight loss, and secondary end points were blood lipid levels and waist-to-hip ratio. Weight loss was significantly greater in the MLC (13.6 lb) than in the NCEP group (7.5 lb), a difference of 6.1 lb (P = .02). There were no significant differences between the groups for total, low density, and high-density lipoprotein cholesterol, triglycerides, or the proportion of small, dense low-density lipoprotein particles. There were significantly favorable changes in all lipid levels within the MLC but not within the NCEP group. Waist-to-hip ratio was not significantly reduced between the groups (P = .27), but it significantly decreased within the MLC group (P = .009). Compared with the NCEP diet, the MLC diet, which is lower in total carbohydrates but higher in complex carbohydrates, protein, and monounsaturated fat, caused significantly greater weight loss over 12 weeks. There were no significant differences between the groups in blood lipid levels, but favorable changes were observed within the MLC diet group.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Zwan, Benjamin J., E-mail: benjamin.zwan@uon.edu.au; O’Connor, Daryl J.; King, Brian W.
2014-08-15
Purpose: To develop a frame-by-frame correction for the energy response of amorphous silicon electronic portal imaging devices (a-Si EPIDs) to radiation that has transmitted through the multileaf collimator (MLC) and to integrate this correction into the backscatter shielded EPID (BSS-EPID) dose-to-water conversion model. Methods: Individual EPID frames were acquired using a Varian frame grabber and iTools acquisition software then processed using in-house software developed inMATLAB. For each EPID image frame, the region below the MLC leaves was identified and all pixels in this region were multiplied by a factor of 1.3 to correct for the under-response of the imager tomore » MLC transmitted radiation. The corrected frames were then summed to form a corrected integrated EPID image. This correction was implemented as an initial step in the BSS-EPID dose-to-water conversion model which was then used to compute dose planes in a water phantom for 35 IMRT fields. The calculated dose planes, with and without the proposed MLC transmission correction, were compared to measurements in solid water using a two-dimensional diode array. Results: It was observed that the integration of the MLC transmission correction into the BSS-EPID dose model improved agreement between modeled and measured dose planes. In particular, the MLC correction produced higher pass rates for almost all Head and Neck fields tested, yielding an average pass rate of 99.8% for 2%/2 mm criteria. A two-sample independentt-test and fisher F-test were used to show that the MLC transmission correction resulted in a statistically significant reduction in the mean and the standard deviation of the gamma values, respectively, to give a more accurate and consistent dose-to-water conversion. Conclusions: The frame-by-frame MLC transmission response correction was shown to improve the accuracy and reduce the variability of the BSS-EPID dose-to-water conversion model. The correction may be applied as a preprocessing step in any pretreatment portal dosimetry calculation and has been shown to be beneficial for highly modulated IMRT fields.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Borges, C.; Zarza-Moreno, M.; Heath, E.
2012-01-15
Purpose: The most recent Varian micro multileaf collimator (MLC), the High Definition (HD120) MLC, was modeled using the BEAMNRC Monte Carlo code. This model was incorporated into a Varian medical linear accelerator, for a 6 MV beam, in static and dynamic mode. The model was validated by comparing simulated profiles with measurements. Methods: The Varian Trilogy (2300C/D) accelerator model was accurately implemented using the state-of-the-art Monte Carlo simulation program BEAMNRC and validated against off-axis and depth dose profiles measured using ionization chambers, by adjusting the energy and the full width at half maximum (FWHM) of the initial electron beam. Themore » HD120 MLC was modeled by developing a new BEAMNRC component module (CM), designated HDMLC, adapting the available DYNVMLC CM and incorporating the specific characteristics of this new micro MLC. The leaf dimensions were provided by the manufacturer. The geometry was visualized by tracing particles through the CM and recording their position when a leaf boundary is crossed. The leaf material density and abutting air gap between leaves were adjusted in order to obtain a good agreement between the simulated leakage profiles and EBT2 film measurements performed in a solid water phantom. To validate the HDMLC implementation, additional MLC static patterns were also simulated and compared to additional measurements. Furthermore, the ability to simulate dynamic MLC fields was implemented in the HDMLC CM. The simulation results of these fields were compared with EBT2 film measurements performed in a solid water phantom. Results: Overall, the discrepancies, with and without MLC, between the opened field simulations and the measurements using ionization chambers in a water phantom, for the off-axis profiles are below 2% and in depth-dose profiles are below 2% after the maximum dose depth and below 4% in the build-up region. On the conditions of these simulations, this tungsten-based MLC has a density of 18.7 g cm{sup -3} and an overall leakage of about 1.1 {+-} 0.03%. The discrepancies between the film measured and simulated closed and blocked fields are below 2% and 8%, respectively. Other measurements were performed for alternated leaf patterns and the agreement is satisfactory (to within 4%). The dynamic mode for this MLC was implemented and the discrepancies between film measurements and simulations are within 4%. Conclusions: The Varian Trilogy (2300 C/D) linear accelerator including the HD120 MLC was successfully modeled and simulated using the Monte Carlo BEAMNRC code by developing an independent CM, the HDMLC CM, either in static and dynamic modes.« less
Falk, Marianne; Larsson, Tobias; Keall, Paul; Chul Cho, Byung; Aznar, Marianne; Korreman, Stine; Poulsen, Per; Munck Af Rosenschold, Per
2012-03-01
Real-time dynamic multileaf collimator (MLC) tracking for management of intrafraction tumor motion can be challenging for highly modulated beams, as the leaves need to travel far to adjust for target motion perpendicular to the leaf travel direction. The plan modulation can be reduced by using a leaf position constraint (LPC) that reduces the difference in the position of adjacent MLC leaves in the plan. The purpose of this study was to investigate the impact of the LPC on the quality of inversely optimized arc radiotherapy plans and the effect of the MLC motion pattern on the dosimetric accuracy of MLC tracking delivery. Specifically, the possibility of predicting the accuracy of MLC tracking delivery based on the plan modulation was investigated. Inversely optimized arc radiotherapy plans were created on CT-data of three lung cancer patients. For each case, five plans with a single 358° arc were generated with LPC priorities of 0 (no LPC), 0.25, 0.5, 0.75, and 1 (highest possible LPC), respectively. All the plans had a prescribed dose of 2 Gy × 30, used 6 MV, a maximum dose rate of 600 MU/min and a collimator angle of 45° or 315°. To quantify the plan modulation, an average adjacent leaf distance (ALD) was calculated by averaging the mean adjacent leaf distance for each control point. The linear relationship between the plan quality [i.e., the calculated dose distributions and the number of monitor units (MU)] and the LPC was investigated, and the linear regression coefficient as well as a two tailed confidence level of 95% was used in the evaluation. The effect of the plan modulation on the performance of MLC tracking was tested by delivering the plans to a cylindrical diode array phantom moving with sinusoidal motion in the superior-inferior direction with a peak-to-peak displacement of 2 cm and a cycle time of 6 s. The delivery was adjusted to the target motion using MLC tracking, guided in real-time by an infrared optical system. The dosimetric results were evaluated using gamma index evaluation with static target measurements as reference. The plan quality parameters did not depend significantly on the LPC (p ≥ 0.066), whereas the ALD depended significantly on the LPC (p < 0.001). The gamma index failure rate depended significantly on the ALD, weighted to the percentage of the beam delivered in each control point of the plan (ALD(w)) when MLC tracking was used (p < 0.001), but not for delivery without MLC tracking (p ≥ 0.342). The gamma index failure rate with the criteria of 2% and 2 mm was decreased from > 33.9% without MLC tracking to <31.4% (LPC 0) and <2.2% (LPC 1) with MLC tracking. The results indicate that the dosimetric robustness of MLC tracking delivery of an inversely optimized arc radiotherapy plan can be improved by incorporating leaf position constraints in the objective function without otherwise affecting the plan quality. The dosimetric robustness may be estimated prior to delivery by evaluating the ALD(w) of the plan.
ERIC Educational Resources Information Center
Kerr, Deirdre; Chung, Gregory K. W. K.; Iseli, Markus R.
2011-01-01
Analyzing log data from educational video games has proven to be a challenging endeavor. In this paper, we examine the feasibility of using cluster analysis to extract information from the log files that is interpretable in both the context of the game and the context of the subject area. If cluster analysis can be used to identify patterns of…
Thrombin-induced activation of RhoA in platelet shape change.
Bodie, S L; Ford, I; Greaves, M; Nixon, G F
2001-09-14
Thrombin-induced activation of RhoA and its involvement in the regulation of myosin II light chain(20) phosphorylation (MLC-P) in alpha-toxin permeabilized platelets was investigated. Permeabilized platelets, expressing normal levels of P-selectin, displayed a Ca(2+)-dependent increase in shape change and MLC-P. Thrombin activated RhoA as measured by a rhotekin-binding assay within 30 s of stimulation under conditions of constant [Ca(2+)](i). Under the same conditions and timecourse, thrombin or GTPgammaS induced an increase in MLC-P and platelet shape change which was not dependent on an increase in [Ca(2+)](i). The thrombin- and GTPgammaS-induced MLC-P in constant [Ca(2+)](i) was inhibited by the addition of Y27632, a Rho-kinase inhibitor. This study directly demonstrates that thrombin can activate RhoA in platelets in a timecourse compatible with a role in increasing MLC-P and shape change (not involving an increase in [Ca(2+)](i)). This is also Rho-kinase-dependent. Copyright 2001 Academic Press.
A virtual source model for Monte Carlo simulation of helical tomotherapy.
Yuan, Jiankui; Rong, Yi; Chen, Quan
2015-01-08
The purpose of this study was to present a Monte Carlo (MC) simulation method based on a virtual source, jaw, and MLC model to calculate dose in patient for helical tomotherapy without the need of calculating phase-space files (PSFs). Current studies on the tomotherapy MC simulation adopt a full MC model, which includes extensive modeling of radiation source, primary and secondary jaws, and multileaf collimator (MLC). In the full MC model, PSFs need to be created at different scoring planes to facilitate the patient dose calculations. In the present work, the virtual source model (VSM) we established was based on the gold standard beam data of a tomotherapy unit, which can be exported from the treatment planning station (TPS). The TPS-generated sinograms were extracted from the archived patient XML (eXtensible Markup Language) files. The fluence map for the MC sampling was created by incorporating the percentage leaf open time (LOT) with leaf filter, jaw penumbra, and leaf latency contained from sinogram files. The VSM was validated for various geometry setups and clinical situations involving heterogeneous media and delivery quality assurance (DQA) cases. An agreement of < 1% was obtained between the measured and simulated results for percent depth doses (PDDs) and open beam profiles for all three jaw settings in the VSM commissioning. The accuracy of the VSM leaf filter model was verified in comparing the measured and simulated results for a Picket Fence pattern. An agreement of < 2% was achieved between the presented VSM and a published full MC model for heterogeneous phantoms. For complex clinical head and neck (HN) cases, the VSM-based MC simulation of DQA plans agreed with the film measurement with 98% of planar dose pixels passing on the 2%/2 mm gamma criteria. For patient treatment plans, results showed comparable dose-volume histograms (DVHs) for planning target volumes (PTVs) and organs at risk (OARs). Deviations observed in this study were consistent with literature. The VSM-based MC simulation approach can be feasibly built from the gold standard beam model of a tomotherapy unit. The accuracy of the VSM was validated against measurements in homogeneous media, as well as published full MC model in heterogeneous media.
A virtual source model for Monte Carlo simulation of helical tomotherapy
Yuan, Jiankui; Rong, Yi
2015-01-01
The purpose of this study was to present a Monte Carlo (MC) simulation method based on a virtual source, jaw, and MLC model to calculate dose in patient for helical tomotherapy without the need of calculating phase‐space files (PSFs). Current studies on the tomotherapy MC simulation adopt a full MC model, which includes extensive modeling of radiation source, primary and secondary jaws, and multileaf collimator (MLC). In the full MC model, PSFs need to be created at different scoring planes to facilitate the patient dose calculations. In the present work, the virtual source model (VSM) we established was based on the gold standard beam data of a tomotherapy unit, which can be exported from the treatment planning station (TPS). The TPS‐generated sinograms were extracted from the archived patient XML (eXtensible Markup Language) files. The fluence map for the MC sampling was created by incorporating the percentage leaf open time (LOT) with leaf filter, jaw penumbra, and leaf latency contained from sinogram files. The VSM was validated for various geometry setups and clinical situations involving heterogeneous media and delivery quality assurance (DQA) cases. An agreement of <1% was obtained between the measured and simulated results for percent depth doses (PDDs) and open beam profiles for all three jaw settings in the VSM commissioning. The accuracy of the VSM leaf filter model was verified in comparing the measured and simulated results for a Picket Fence pattern. An agreement of <2% was achieved between the presented VSM and a published full MC model for heterogeneous phantoms. For complex clinical head and neck (HN) cases, the VSM‐based MC simulation of DQA plans agreed with the film measurement with 98% of planar dose pixels passing on the 2%/2 mm gamma criteria. For patient treatment plans, results showed comparable dose‐volume histograms (DVHs) for planning target volumes (PTVs) and organs at risk (OARs). Deviations observed in this study were consistent with literature. The VSM‐based MC simulation approach can be feasibly built from the gold standard beam model of a tomotherapy unit. The accuracy of the VSM was validated against measurements in homogeneous media, as well as published full MC model in heterogeneous media. PACS numbers: 87.53.‐j, 87.55.K‐ PMID:25679157
Code of Federal Regulations, 2013 CFR
2013-04-01
... include staff time associated with: (A) Processing FOIA requests; (B) Locating and reviewing files; (C) Monitoring file reviews; (D) Generating computer records (electronic print-outs); and (E) Preparing logs of..., black and white copies. The charge for copying standard sized, black and white public records shall be...
Code of Federal Regulations, 2012 CFR
2012-04-01
... include staff time associated with: (A) Processing FOIA requests; (B) Locating and reviewing files; (C) Monitoring file reviews; (D) Generating computer records (electronic print-outs); and (E) Preparing logs of..., black and white copies. The charge for copying standard sized, black and white public records shall be...
Poster - 53: Improving inter-linac DMLC IMRT dose precision by fine tuning of MLC leaf calibration
DOE Office of Scientific and Technical Information (OSTI.GOV)
Nakonechny, Keith; Tran, Muoi; Sasaki, David
Purpose: To develop a method to improve the inter-linac precision of DMLC IMRT dosimetry. Methods: The distance between opposing MLC leaf banks (“gap size”) can be finely tuned on Varian linacs. The dosimetric effect due to small deviations from the nominal gap size (“gap error”) was studied by introducing known errors for several DMLC sliding gap sizes, and for clinical plans based on the TG119 test cases. The plans were delivered on a single Varian linac and the relationship between gap error and the corresponding change in dose was measured. The plans were also delivered on eight Varian 2100 seriesmore » linacs (at two institutions) in order to quantify the inter-linac variation in dose before and after fine tuning the MLC calibration. Results: The measured dose differences for each field agreed well with the predictions of LoSasso et al. Using the default MLC calibration, the variation in the physical MLC gap size was determined to be less than 0.4 mm between all linacs studied. The dose difference between the linacs with the largest and smallest physical gap was up to 5.4% (spinal cord region of the head and neck TG119 test case). This difference was reduced to 2.5% after fine tuning the MLC gap calibration. Conclusions: The inter-linac dose precision for DMLC IMRT on Varian linacs can be improved using a simple modification of the MLC calibration procedure that involves fine adjustment of the nominal gap size.« less
Minamisawa, S; Gu, Y; Ross, J; Chien, K R; Chen, J
1999-04-09
Our previous study of homozygous mutants of the ventricular specific isoform of myosin light chain 2 (mlc-2v) demonstrated that mlc-2v plays an essential role in murine heart development (Chen, J., Kubalak, S. W., Minamisawa, S., Price, R. L., Becker, K. D., Hickey, R., Ross, J., Jr., and Chien, K. R. (1998) J. Biol. Chem. 273, 1252-1256). As gene dosage of some myofibrillar proteins can affect muscle function, we have analyzed heterozygous mutants in depth. Ventricles of heterozygous mutants displayed a 50% reduction in mlc-2v mRNA, yet expressed normal levels of protein both under basal conditions and following induction of cardiac hypertrophy by aortic constriction. Heterozygous mutants exhibited cardiac function comparable to that of wild-type littermate controls both prior to and following aortic constriction. There were no significant differences in contractility and responses to calcium between wild-type and heterozygous unloaded cardiomyocytes. We conclude that heterozygous mutants show neither a molecular nor a physiological cardiac phenotype either at base line or following hypertrophic stimuli. These results suggest that post-transcriptional compensatory mechanisms play a major role in maintaining the level of MLC-2v protein in murine hearts. In addition, as our mlc-2v knockout mutants were created by a knock-in of Cre recombinase into the endogenous mlc-2v locus, this study demonstrates that heterozygous mlc-2v cre knock-in mice are appropriate for ventricular specific gene targeting.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Otageri, P; Grant, E; Maricle, S
Purpose: To evaluate the effects of MLC modeling after commissioning the Varian TrueBeam LINAC in Pinnacle version 9.2. Methods: Stepand-shoot IMRT QAs were investigated when we observed our measured absolute dose results using ion chamber (Capintec PR-05P) were uncharacteristically low; about 4–5% compared to doses calculated by Pinnacle{sup 3} (Phillips, Madison, WI). This problem was predominant for large and highly modulated head and neck (HN) treatments. Intuitively we knew this had to be related to shortcomings in the MLC modeling in Pinnacle. Using film QA we were able to iteratively adjust the MLC parameters. We confirmed results by re-testing fivemore » failed IMRT QA patients; and ion chamber measurements were verified in Quasar anthropomorphic phantom. Results: After commissioning the LINAC in Pinnacle version 9.2, the MLC transmission for 6X, 10X and 15X were 2.0%, 1.7% and 2.0%, respectively, and additional Interleaf leakage for all three energies was 0.5%. These parameters were obtained from profiles scanned with an Edge detector (Sun Nuclear, Melbourne, FL) during machine commissioning. A Verification testing with radiographic EDR2 film (Kodak, Rochester, NY) measurement was performed by creating a closed MLC leaf pattern and analyzing using RIT software (RIT, Colorado Springs, CO). This reduced MLC transmission for 6X, 10X and 15X to 0.7%, 0.9% and 0.9%, respectively; while increasing additional Interleaf leakage for all three energies to 1.0%. Conclusion: Radiographic film measurements were used to correct MLC transmission values for step and shoot IMRT fields used in Pinnacle version 9.2. After adjusting the MLC parameters to correlate with the film QA, there was still very good agreement between the Pinnacle model and commissioning data. Using the same QA methodology, we were also able to improve the beam models for the Varian C-series linacs, Novalis-Tx, and TrueBeam M-120 linacs.« less
SU-E-T-195: Gantry Angle Dependency of MLC Leaf Position Error
DOE Office of Scientific and Technical Information (OSTI.GOV)
Ju, S; Hong, C; Kim, M
Purpose: The aim of this study was to investigate the gantry angle dependency of the multileaf collimator (MLC) leaf position error. Methods: An automatic MLC quality assurance system (AutoMLCQA) was developed to evaluate the gantry angle dependency of the MLC leaf position error using an electronic portal imaging device (EPID). To eliminate the EPID position error due to gantry rotation, we designed a reference maker (RM) that could be inserted into the wedge mount. After setting up the EPID, a reference image was taken of the RM using an open field. Next, an EPID-based picket-fence test (PFT) was performed withoutmore » the RM. These procedures were repeated at every 45° intervals of the gantry angle. A total of eight reference images and PFT image sets were analyzed using in-house software. The average MLC leaf position error was calculated at five pickets (-10, -5, 0, 5, and 10 cm) in accordance with general PFT guidelines using in-house software. This test was carried out for four linear accelerators. Results: The average MLC leaf position errors were within the set criterion of <1 mm (actual errors ranged from -0.7 to 0.8 mm) for all gantry angles, but significant gantry angle dependency was observed in all machines. The error was smaller at a gantry angle of 0° but increased toward the positive direction with gantry angle increments in the clockwise direction. The error reached a maximum value at a gantry angle of 90° and then gradually decreased until 180°. In the counter-clockwise rotation of the gantry, the same pattern of error was observed but the error increased in the negative direction. Conclusion: The AutoMLCQA system was useful to evaluate the MLC leaf position error for various gantry angles without the EPID position error. The Gantry angle dependency should be considered during MLC leaf position error analysis.« less
Uematsu, Masahiro; Ito, Makiko; Hama, Yukihiro; Inomata, Takayuki; Fujii, Masahiro; Nishio, Teiji; Nakamura, Naoki; Nakagawa, Keiichi
2012-01-01
In this paper, we suggest a new method for verifying the motion of a binary multileaf collimator (MLC) in helical tomotherapy. For this we used a combination of a cylindrical scintillator and a general‐purpose camcorder. The camcorder records the light from the scintillator following photon irradiation, which we use to track the motion of the binary MLC. The purpose of this study is to demonstrate the feasibility of this method as a binary MLC quality assurance (QA) tool. First, the verification was performed using a simple binary MLC pattern with a constant leaf open time; secondly, verification using the binary MLC pattern used in a clinical setting was also performed. Sinograms of simple binary MLC patterns, in which leaves that were open were detected as “open” from the measured light, define the sensitivity which, in this case, was 1.000. On the other hand, the specificity, which gives the fraction of closed leaves detected as “closed”, was 0.919. The leaf open error identified by our method was −1.3±7.5%. The 68.6% of observed leaves were performed within ± 3% relative error. The leaf open error was expressed by the relative errors calculated on the sinogram. In the clinical binary MLC pattern, the sensitivity and specificity were 0.994 and 0.997, respectively. The measurement could be performed with −3.4±8.0% leaf open error. The 77.5% of observed leaves were performed within ± 3% relative error. With this method, we can easily verify the motion of the binary MLC, and the measurement unit developed was found to be an effective QA tool. PACS numbers: 87.56.Fc, 87.56.nk PMID:22231222
SU-F-T-372: Surface and Peripheral Dose in Compensator-Based FFF Beam IMRT
DOE Office of Scientific and Technical Information (OSTI.GOV)
Zhang, D; Feygelman, V; Moros, E
2016-06-15
Purpose: Flattening filter free (FFF) beams produce higher dose rates. Combined with compensator IMRT techniques, the dose delivery for each beam can be much shorter compared to the flattened beam MLC-based or compensator-based IMRT. This ‘snap shot’ IMRT delivery is beneficial to patients for tumor motion management. Due to softer energy, surface doses in FFF beam treatment are usually higher than those from flattened beams. Because of less scattering due to no flattening filter, peripheral doses are usually lower in FFF beam treatment. However, in compensator-based IMRT using FFF beams, the compensator is in the beam pathway. Does it introducemore » beam hardening effects and scattering such that the surface dose is lower and peripheral dose is higher compared to FFF beam MLC-based IMRT? Methods: This study applied Monte Carlo techniques to investigate the surface and peripheral doses in compensator-based IMRT using FFF beams and compared it to the MLC-based IMRT using FFF beams and flattened beams. Besides various thicknesses of copper slabs to simulate various thicknesses of compensators, a simple cone-shaped compensator was simulated to mimic a clinical application. The dose distribution in water phantom by the cone-shaped compensator was then simulated by multiple MLC defined FFF and flattened beams with various openings. After normalized to Dmax, the surface and peripheral dose was compared between the FFF beam compensator-based IMRT and FFF/flattened beam MLC-based IMRT. Results: The surface dose at the central 0.5mm depth was close between the compensator and 6FFF MLC dose distributions, and about 8% (of Dmax) higher than the flattened 6MV MLC dose. At 8cm off axis at dmax, the peripheral dose between the 6FFF and flattened 6MV MLC demonstrated similar doses, while the compensator dose was about 1% higher. Conclusion: Compensator does not reduce the surface doses but slightly increases the peripheral doses due to scatter inside compensator.« less
Visser, R; Godart, J; Wauben, D J L; Langendijk, J A; Van't Veld, A A; Korevaar, E W
2016-05-21
The objective of this study was to introduce a new iterative method to reconstruct multi leaf collimator (MLC) positions based on low resolution ionization detector array measurements and to evaluate its error detection performance. The iterative reconstruction method consists of a fluence model, a detector model and an optimizer. Expected detector response was calculated using a radiotherapy treatment plan in combination with the fluence model and detector model. MLC leaf positions were reconstructed by minimizing differences between expected and measured detector response. The iterative reconstruction method was evaluated for an Elekta SLi with 10.0 mm MLC leafs in combination with the COMPASS system and the MatriXX Evolution (IBA Dosimetry) detector with a spacing of 7.62 mm. The detector was positioned in such a way that each leaf pair of the MLC was aligned with one row of ionization chambers. Known leaf displacements were introduced in various field geometries ranging from -10.0 mm to 10.0 mm. Error detection performance was tested for MLC leaf position dependency relative to the detector position, gantry angle dependency, monitor unit dependency, and for ten clinical intensity modulated radiotherapy (IMRT) treatment beams. For one clinical head and neck IMRT treatment beam, influence of the iterative reconstruction method on existing 3D dose reconstruction artifacts was evaluated. The described iterative reconstruction method was capable of individual MLC leaf position reconstruction with millimeter accuracy, independent of the relative detector position within the range of clinically applied MU's for IMRT. Dose reconstruction artifacts in a clinical IMRT treatment beam were considerably reduced as compared to the current dose verification procedure. The iterative reconstruction method allows high accuracy 3D dose verification by including actual MLC leaf positions reconstructed from low resolution 2D measurements.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Vacca, N; Caussa, L; Filipuzzi, M
2014-06-01
Purpose: The purpose of this work was to evaluate the dosimetric characteristics of three collimation systems, 5mm circular cone (Brainlab) and square fields of 5mm with HDMLC (Varian) and microMLC Moduleaf, Siemens) for trigeminal neuralgia treatment. Methods: A TPS Iplan v4.5 BrainLAB was used to do treatment plans for each collimations system in a square solid water phantom with isocenter at 5cm depth. Single field and treatment plan including 11 arcs with fix field and 100° gantry range was made for each collimation systems. EBT3 films were positioned at isocenter in a coronal plane to measured dose distribution for allmore » geometries. Films were digitized with a Vidar DosimetryPro Red scanner with a resolution of 89dpi and RIT113v6.1 software was used for analysis. Penumbra region (80%–20%), FWHM and dose percentage at 5mm and 10mm from CAX were determined. All profiles were normalized at CAX. Results: For single beam the penumbra (FWHM) was 1.5mm (5.3mm) for the cone, 1.9mm (5.5mm) for HDMLC and 1.8mm (5.4mm) for the microMLC. Dose percentage at 5mm was 6.9% for cone, 12.5% for HDMLC and 8.7% for the microMLC. For treatment plan the penumbra (FWHM) was 2.58mm (5.47mm) for the cone, 2.8mm (5.84mm) for HDMLC and 2.58mm (6.09mm) for the microMLC. Dose perecentage at 5mm was 13.1% for cone, 16.1% for HDMLC, 15.2% for the microMLC. Conclusion: The cone has a dose falloff larger than the microMLC and HDMLC, by its reduced penumbra, this translates into better protection of surrounding healthy tissue, however, the microMLC and HDMLC have similar accuracy to cone.« less
Electron beam collimation with a photon MLC for standard electron treatments
NASA Astrophysics Data System (ADS)
Mueller, S.; Fix, M. K.; Henzen, D.; Frei, D.; Frauchiger, D.; Loessl, K.; Stampanoni, M. F. M.; Manser, P.
2018-01-01
Standard electron treatments are currently still performed using standard or molded patient-specific cut-outs placed in the electron applicator. Replacing cut-outs and electron applicators with a photon multileaf collimator (pMLC) for electron beam collimation would make standard electron treatments more efficient and would facilitate advanced treatment techniques like modulated electron radiotherapy (MERT) and mixed beam radiotherapy (MBRT). In this work, a multiple source Monte Carlo beam model for pMLC shaped electron beams commissioned at a source-to-surface distance (SSD) of 70 cm is extended for SSDs of up to 100 cm and validated for several Varian treatment units with field sizes typically used for standard electron treatments. Measurements and dose calculations agree generally within 3% of the maximal dose or 2 mm distance to agreement. To evaluate the dosimetric consequences of using pMLC collimated electron beams for standard electron treatments, pMLC-based and cut-out-based treatment plans are created for a left and a right breast boost, a sternum, a testis and a parotid gland case. The treatment plans consist of a single electron field, either alone (1E) or in combination with two 3D conformal tangential photon fields (1E2X). For each case, a pMLC plan with similar treatment plan quality in terms of dose homogeneity to the target and absolute mean dose values to the organs at risk (OARs) compared to a cut-out plan is found. The absolute mean dose to an OAR is slightly increased for pMLC-based compared to cut-out-based 1E plans if the OAR is located laterally close to the target with respect to beam direction, or if a 6 MeV electron beam is used at an extended SSD. In conclusion, treatment plans using cut-out collimation can be replaced by plans of similar treatment plan quality using pMLC collimation with accurately calculated dose distributions.
Wang, Iris Z.; Kumaraswamy, Lalith K.; Podgorsak, Matthew B.
2016-01-01
Background This study is to report 1) the sensitivity of intensity modulated radiation therapy (IMRT) QA method for clinical volumetric modulated arc therapy (VMAT) plans with multi-leaf collimator (MLC) leaf errors that will not trigger MLC interlock during beam delivery; 2) the effect of non-beam-hold MLC leaf errors on the quality of VMAT plan dose delivery. Materials and methods. Eleven VMAT plans were selected and modified using an in-house developed software. For each control point of a VMAT arc, MLC leaves with the highest speed (1.87-1.95 cm/s) were set to move at the maximal allowable speed (2.3 cm/s), which resulted in a leaf position difference of less than 2 mm. The modified plans were considered as ‘standard’ plans, and the original plans were treated as the ‘slowing MLC’ plans for simulating ‘standard’ plans with leaves moving at relatively lower speed. The measurement of each ‘slowing MLC’ plan using MapCHECK®2 was compared with calculated planar dose of the ‘standard’ plan with respect to absolute dose Van Dyk distance-to-agreement (DTA) comparisons using 3%/3 mm and 2%/2 mm criteria. Results All ‘slowing MLC’ plans passed the 90% pass rate threshold using 3%/3 mm criteria while one brain and three anal VMAT cases were below 90% with 2%/2 mm criteria. For ten out of eleven cases, DVH comparisons between ‘standard’ and ‘slowing MLC’ plans demonstrated minimal dosimetric changes in targets and organs-at-risk. Conclusions For highly modulated VMAT plans, pass rate threshold (90%) using 3%/3mm criteria is not sensitive in detecting MLC leaf errors that will not trigger the MLC leaf interlock. However, the consequential effects of non-beam hold MLC errors on target and OAR doses are negligible, which supports the reliability of current patient-specific IMRT quality assurance (QA) method for VMAT plans. PMID:27069458
Modeling of a multileaf collimator
NASA Astrophysics Data System (ADS)
Kim, Siyong
A comprehensive physics model of a multileaf collimator (MLC) field for treatment planning was developed. Specifically, an MLC user interface module that includes a geometric optimization tool and a general method of in- air output factor calculation were developed. An automatic tool for optimization of MLC conformation is needed to realize the potential benefits of MLC. It is also necessary that a radiation therapy treatment planning (RTTP) system is capable of modeling MLC completely. An MLC geometric optimization and user interface module was developed. The planning time has been reduced significantly by incorporating the MLC module into the main RTTP system, Radiation Oncology Computer System (ROCS). The dosimetric parameter that has the most profound effect on the accuracy of the dose delivered with an MLC is the change in the in-air output factor that occurs with field shaping. It has been reported that the conventional method of calculating an in-air output factor cannot be used for MLC shaped fields accurately. Therefore, it is necessary to develop algorithms that allow accurate calculation of the in-air output factor. A generalized solution for an in-air output factor calculation was developed. Three major contributors of scatter to the in-air output-flattening filter, wedge, and tertiary collimator-were considered separately. By virtue of a field mapping method, in which a source plane field determined by detector's eye view is mapped into a detector plane field, no additional dosimetric data acquisition other than the standard data set for a range of square fields is required for the calculation of head scatter. Comparisons of in-air output factors between calculated and measured values show a good agreement for both open and wedge fields. For rectangular fields, a simple equivalent square formula was derived based on the configuration of a linear accelerator treatment head. This method predicts in-air output to within 1% accuracy. A two-effective-source algorithm was developed to account for the effect of source to detector distance on in-air output for wedge fields. Two effective sources, one for head scatter and the other for wedge scatter, were dealt with independently. Calculations provided less than 1% difference of in-air output factors from measurements. This approach offers the best comprehensive accuracy in radiation delivery with field shapes defined using MLC. This generalized model works equally well with fields shaped by any type of tertiary collimator and have the necessary framework to extend its application to intensity modulated radiation therapy.
9 CFR 327.10 - Samples; inspection of consignments; refusal of entry; marking.
Code of Federal Regulations, 2010 CFR
2010-01-01
... import establishment and approved by the Director, Import Inspection Division, is on file at the import... (iv) That the establishment will maintain a daily stamping log containing the following information... covering the product to be inspected. The daily stamping log must be retained by the establishment in...
46 CFR 78.37-3 - Logbooks and records.
Code of Federal Regulations, 2011 CFR
2011-10-01
... completed, the master or person in charge shall file the logbook with the Officer in Charge, Marine... purposes of making entries therein as required by law or regulations in this subchapter. Such logs or... records of tests and inspections of fire fighting equipment must be maintained with the vessel's logs for...
46 CFR 131.610 - Logbooks and records.
Code of Federal Regulations, 2011 CFR
2011-10-01
... COAST GUARD, DEPARTMENT OF HOMELAND SECURITY (CONTINUED) OFFSHORE SUPPLY VESSELS OPERATIONS Logs § 131... them. (d) When a voyage is completed, or after a specified time has elapsed, the master shall file the... alternative log or record for making entries required by law, including regulations in this subchapter. This...
46 CFR 131.610 - Logbooks and records.
Code of Federal Regulations, 2010 CFR
2010-10-01
... COAST GUARD, DEPARTMENT OF HOMELAND SECURITY (CONTINUED) OFFSHORE SUPPLY VESSELS OPERATIONS Logs § 131... them. (d) When a voyage is completed, or after a specified time has elapsed, the master shall file the... alternative log or record for making entries required by law, including regulations in this subchapter. This...
9 CFR 327.10 - Samples; inspection of consignments; refusal of entry; marking.
Code of Federal Regulations, 2011 CFR
2011-01-01
... import establishment and approved by the Director, Import Inspection Division, is on file at the import... (iv) That the establishment will maintain a daily stamping log containing the following information... covering the product to be inspected. The daily stamping log must be retained by the establishment in...
46 CFR 78.37-3 - Logbooks and records.
Code of Federal Regulations, 2010 CFR
2010-10-01
... completed, the master or person in charge shall file the logbook with the Officer in Charge, Marine... purposes of making entries therein as required by law or regulations in this subchapter. Such logs or... records of tests and inspections of fire fighting equipment must be maintained with the vessel's logs for...
Code of Federal Regulations, 2014 CFR
2014-07-01
... test to generate a submission package file, which documents performance test data. You must then submit the file generated by the ERT through the EPA's Compliance and Emissions Data Reporting Interface (CEDRI), which can be accessed by logging in to the EPA's Central Data Exchange (CDX) (https://cdx.epa...
TM 4: Beam through the Main Linac Cryomodule
DOE Office of Scientific and Technical Information (OSTI.GOV)
Bartnik, A.
2017-06-14
On May 15th 2017, the CBETA project reached the major funding milestone, “Beam through the MLC.” For this test, the team had to successfully accelerate the electron beam to 6 MeV in the Injector Cryomodule (ICM), and then to a final energy of 12 MeV in the Main Linac Cryomodule (MLC). The MLC contains six superconducting accelerating cavities; for this initial test only a single cavity was powered.
Chen, Jiayun; Fu, Guishan; Li, Minghui; Song, Yixin; Dai, Jianrong; Miao, Junjie; Liu, Zhiqiang; Li, Yexiong
2017-12-14
The purpose of this paper was to evaluate the impact of leaf treatment of multileaf collimator (MLC) in plan quality of intensity-modulated radiotherapy (IMRT) of patients with advanced lung cancer. Five MLCs with different leaf transmissions (0.01%, 0.5%, 1.2%, 1.8%, and 3%) were configured for an accelerator in a treatment planning system. Correspondingly, 5 treatment plans with the same optimization setting were created and evaluated quantitatively for each patient (11 patients total) who was diagnosed with advanced lung cancer. All of the 5 plans for each patient met the dose requirement for the planning treatment volumes (PTVs) and had similar target dose homogeneity and conformity. On average, the doses to selected organs were as follows: (1) V 5 , V 20 , and the mean dose of total lung; (2) the maximum and mean dose to spinal cord planning organ-at-risk volume (PRV); and (3) V 30 and V 40 of heart, decreased slightly when MLC transmission was decreased, but with no statistical differences. There is a clear grouping of plans having total quality score (S D ) value, which is used to evaluate plan quality: (1) more than 1 (patient nos. 1 to 3, 5, and 8), and more than 2.5 (patient no. 6); (2) less than 1 (patient nos. 7 and 10); (3) around 1 (patient nos. 4, 9, and 11). As MLC transmission increased, overall S D values increased as well and plan dose requirement was harder to meet. The clinical requirements were violated increasingly as MLC transmission became large. Total S D with and without normal tissue (NT) showed similar results, with no statistically significant differences. Therefore, decrease of MLC transmission did have minimum impact on plan, and it improved target coverage and reduced normal tissue radiation slightly, with no statistical significance. Plan quality could not be significantly improved by MLC transmission reduction. However, lower MLC transmission may have advantages on lung sparing to low- and intermediate-dose exposure. Besides conventional fraction, hyperfraction, or stereotactic body radiotherapy (SBRT), the reduction on lung sparing is still essential because it is highly relevant to radiation pneumonitis (RP). It has potential to diminish incidence of RP and improve patient's quality of life after irradiation with lowered MLC transmission. Copyright © 2017 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Han, B; Xing, L; Wang, L
Purpose: To systematically investigate an ultra-high spatial-resolution amorphous silicon flat-panel electronic portal imaging device (EPID) for MLC-based full-body robotic radiosurgery geometric and dosimetric quality assurance (QA). Methods: The high frame-rate and ultra-high spatial resolution EPID is an outstanding detector for measuring profiles, MLC-shaped radiosurgery field aperture verification, and small field dosimetry. A Monte Carlo based technique with a robotic linac specific response and calibration is developed to convert a raw EPID-measured image of a radiosurgery field into water-based dose distribution. The technique is applied to measure output factors and profiles for 6MV MLC-defined radiosurgery fields with various sizes ranging frommore » 7.6mm×7.7mm to 100mm×100.1mm and the results are compared with the radiosurgery diode scan measurements in water tank. The EPID measured field sizes and the penumbra regions are analyzed to evaluate the MLC positioning accuracy. Results: For all MLC fields, the EPID measured output factors of MLC-shaped fields are in good agreement with the diode measurements. The mean output difference between the EPID and diode measurement is 0.05±0.87%. The max difference is −1.33% for 7.6mm×7.7mm field. The MLC field size derived from the EPID measurements are in good agreement comparing to the diode scan result. For crossline field sizes, the mean difference is −0.17mm±0.14mm with a maximum of −0.35mm for the 30.8mm×30.8mm field. For inline field sizes, the mean difference is +0.08mm±0.18mm with a maximum of +0.45mm for the 100mm×100.1mm field. The high resolution EPID is able to measure the whole radiation field, without the need to align the detector center perfectly at field center as diode or ion chamber measurement. The setup time is greatly reduced so that the whole process is possible for machine and patient-specific QA. Conclusion: The high spatial-resolution EPID is proved to be an accurate and efficient tool for QA of MLC-equipped robotic radiosurgery system.« less
Developing a Complete and Effective ACT-R Architecture
2008-01-01
of computational primitives , as contrasted with the predominant “one-off” and “grab-bag” cognitive models in the field. These architectures have...transport/ semaphore protocols connected via a glue script. Both protocols rely on the fact that file rename and file remove operations are atomic...the Trial Log file until just prior to processing the next input request. Thus, to perform synchronous identifications it is necessary to run an
2010-01-01
Background Regulatory elements that control expression of specific genes during development have been shown in many cases to contain functionally-conserved modules that can be transferred between species and direct gene expression in a comparable developmental pattern. An example of such a module has been identified at the rat myosin light chain (MLC) 1/3 locus, which has been well characterised in transgenic mouse studies. This locus contains two promoters encoding two alternatively spliced isoforms of alkali myosin light chain. These promoters are differentially regulated during development through the activity of two enhancer elements. The MLC3 promoter alone has been shown to confer expression of a reporter gene in skeletal and cardiac muscle in transgenic mice and the addition of the downstream MLC enhancer increased expression levels in skeletal muscle. We asked whether this regulatory module, sufficient for striated muscle gene expression in the mouse, would drive expression in similar domains in the chicken. Results We have observed that a conserved downstream MLC enhancer is present in the chicken MLC locus. We found that the rat MLC1/3 regulatory elements were transcriptionally active in chick skeletal muscle primary cultures. We observed that a single copy lentiviral insert containing this regulatory cassette was able to drive expression of a lacZ reporter gene in the fast-fibres of skeletal muscle in chicken in three independent transgenic chicken lines in a pattern similar to the endogenous MLC locus. Reporter gene expression in cardiac muscle tissues was not observed for any of these lines. Conclusions From these results we conclude that skeletal expression from this regulatory module is conserved in a genomic context between rodents and chickens. This transgenic module will be useful in future investigations of muscle development in avian species. PMID:20184756
Techtalk: Telecommunications for Improving Developmental Education.
ERIC Educational Resources Information Center
Caverly, David C.; Broderick, Bill
1993-01-01
Explains how to access the Internet, discussing hardware and software considerations, connectivity, and types of access available to users. Describes the uses of electronic mail; TELNET, a method for remotely logging onto another computer; and anonymous File Transfer Protocol (FTP), a method for downloading files from a remote computer. (MAB)
Politi, Liran; Codish, Shlomi; Sagy, Iftach; Fink, Lior
2014-12-01
Insights about patterns of system use are often gained through the analysis of system log files, which record the actual behavior of users. In a clinical context, however, few attempts have been made to typify system use through log file analysis. The present study offers a framework for identifying, describing, and discerning among patterns of use of a clinical information retrieval system. We use the session attributes of volume, diversity, granularity, duration, and content to define a multidimensional space in which each specific session can be positioned. We also describe an analytical method for identifying the common archetypes of system use in this multidimensional space. We demonstrate the value of the proposed framework with a log file of the use of a health information exchange (HIE) system by physicians in an emergency department (ED) of a large Israeli hospital. The analysis reveals five distinct patterns of system use, which have yet to be described in the relevant literature. The results of this study have the potential to inform the design of HIE systems for efficient and effective use, thus increasing their contribution to the clinical decision-making process. Copyright © 2014 Elsevier Inc. All rights reserved.
Myosin light chain 2-based selection of human iPSC-derived early ventricular cardiac myocytes
Bizy, Alexandra; Guerrero-Serna, Guadalupe; Hu, Bin; Ponce-Balbuena, Daniela; Willis, B. Cicero; Zarzoso, Manuel; Ramirez, Rafael J.; Sener, Michelle F.; Mundada, Lakshmi V.; Klos, Matthew; Devaney, Eric J.; Vikstrom, Karen L.; Herron, Todd J.; Jalife, José
2014-01-01
Applications of human induced pluripotent stemcell derived-cardiac myocytes (hiPSC-CMs) would be strengthened by the ability to generate specific cardiac myocyte (CM) lineages. However, purification of lineage-specific hiPSC-CMs is limited by the lack of cell marking techniques. Here, we have developed an iPSC-CM marking system using recombinant adenoviral reporter constructs with atrial- or ventricular-specific myosin light chain-2 (MLC-2) promoters. MLC-2a and MLC-2v selected hiPSC-CMs were purified by fluorescence-activated cell sorting and their biochemical and electrophysiological phenotypes analyzed. We demonstrate that the phenotype of both populations remained stable in culture and they expressed the expected sarcomeric proteins, gap junction proteins and chamber-specific transcription factors. Compared to MLC-2a cells, MLC-2v selected CMs had larger action potential amplitudes and durations. In addition, by immunofluorescence, we showed that MLC-2 isoform expression can be used to enrich hiPSC-CM consistent with early atrial and ventricularmyocyte lineages. However, only the ventricular myosin light chain-2 promoter was able to purify a highly homogeneous population of iPSC-CMs. Using this approach, it is now possible to develop ventricular-specific disease models using iPSC-CMs while atrial-specific iPSC-CM cultures may require additional chamber-specific markers. PMID:24095945
Hey, Jennifer; Roberts, Keith John; Morris-Stiff, Gareth J; Toogood, Giles J
2012-01-01
Objectives Single-incision laparoscopic cholecystectomy (SILC) may be associated with less pain, shorter hospital stay and better cosmetic results than multiport laparoscopic cholecystectomy (MLC). Advocates suggest that patients prefer SILC, although research directly addressing the question of patient preferences is limited. This study aimed to assess patient preferences using currently available evidence. Methods Patients awaiting elective cholecystectomy were shown a series of postoperative images taken after SILC or MLC and asked which procedure this led them to prefer. This was repeated after patients had completed a questionnaire constructed using published objective data comparing patient-reported outcomes of SILC and MLC. Results The study was completed by 113 consecutive patients. After their initial viewing of the images, 16% of subjects preferred MLC. Younger age, lower body mass index and female sex were associated with choosing SILC. After completing the questionnaire, 88% of patients preferred MLC (P < 0.001). Patients ranked the level of risk for complications and postoperative pain above cosmetic results in determining their choice of procedure. Conclusions Patients' initial preference when presented with cosmetic appearance was for SILC. When contemporary outcome data were included, the majority chose MLC. This underlines the need to fully inform patients during the consent process and indicates that patient views of SILC may differ from the views of those introducing the technology. PMID:22404262
Myosin light chain 2-based selection of human iPSC-derived early ventricular cardiac myocytes.
Bizy, Alexandra; Guerrero-Serna, Guadalupe; Hu, Bin; Ponce-Balbuena, Daniela; Willis, B Cicero; Zarzoso, Manuel; Ramirez, Rafael J; Sener, Michelle F; Mundada, Lakshmi V; Klos, Matthew; Devaney, Eric J; Vikstrom, Karen L; Herron, Todd J; Jalife, José
2013-11-01
Applications of human induced pluripotent stem cell derived-cardiac myocytes (hiPSC-CMs) would be strengthened by the ability to generate specific cardiac myocyte (CM) lineages. However, purification of lineage-specific hiPSC-CMs is limited by the lack of cell marking techniques. Here, we have developed an iPSC-CM marking system using recombinant adenoviral reporter constructs with atrial- or ventricular-specific myosin light chain-2 (MLC-2) promoters. MLC-2a and MLC-2v selected hiPSC-CMs were purified by fluorescence-activated cell sorting and their biochemical and electrophysiological phenotypes analyzed. We demonstrate that the phenotype of both populations remained stable in culture and they expressed the expected sarcomeric proteins, gap junction proteins and chamber-specific transcription factors. Compared to MLC-2a cells, MLC-2v selected CMs had larger action potential amplitudes and durations. In addition, by immunofluorescence, we showed that MLC-2 isoform expression can be used to enrich hiPSC-CM consistent with early atrial and ventricular myocyte lineages. However, only the ventricular myosin light chain-2 promoter was able to purify a highly homogeneous population of iPSC-CMs. Using this approach, it is now possible to develop ventricular-specific disease models using iPSC-CMs while atrial-specific iPSC-CM cultures may require additional chamber-specific markers. © 2013.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Ilsche, Thomas; Schuchart, Joseph; Cope, Joseph
Event tracing is an important tool for understanding the performance of parallel applications. As concurrency increases in leadership-class computing systems, the quantity of performance log data can overload the parallel file system, perturbing the application being observed. In this work we present a solution for event tracing at leadership scales. We enhance the I/O forwarding system software to aggregate and reorganize log data prior to writing to the storage system, significantly reducing the burden on the underlying file system for this type of traffic. Furthermore, we augment the I/O forwarding system with a write buffering capability to limit the impactmore » of artificial perturbations from log data accesses on traced applications. To validate the approach, we modify the Vampir tracing tool to take advantage of this new capability and show that the approach increases the maximum traced application size by a factor of 5x to more than 200,000 processors.« less
Fast regional readout CMOS Image Sensor for dynamic MLC tracking
NASA Astrophysics Data System (ADS)
Zin, H.; Harris, E.; Osmond, J.; Evans, P.
2014-03-01
Advanced radiotherapy techniques such as volumetric modulated arc therapy (VMAT) require verification of the complex beam delivery including tracking of multileaf collimators (MLC) and monitoring the dose rate. This work explores the feasibility of a prototype Complementary metal-oxide semiconductor Image Sensor (CIS) for tracking these complex treatments by utilising fast, region of interest (ROI) read out functionality. An automatic edge tracking algorithm was used to locate the MLC leaves edges moving at various speeds (from a moving triangle field shape) and imaged with various sensor frame rates. The CIS demonstrates successful edge detection of the dynamic MLC motion within accuracy of 1.0 mm. This demonstrates the feasibility of the sensor to verify treatment delivery involving dynamic MLC up to ~400 frames per second (equivalent to the linac pulse rate), which is superior to any current techniques such as using electronic portal imaging devices (EPID). CIS provides the basis to an essential real-time verification tool, useful in accessing accurate delivery of complex high energy radiation to the tumour and ultimately to achieve better cure rates for cancer patients.
46 CFR 196.35-3 - Logbooks and records.
Code of Federal Regulations, 2010 CFR
2010-10-01
... form CG-706 or in the owner's format for an official logbook. Such logs must be kept available for a... master or person in charge shall file the logbook with the Officer in Charge, Marine Inspection. (b) The... of making entries therein as required by law or regulations in this subchapter. Such logs or records...
46 CFR 196.35-3 - Logbooks and records.
Code of Federal Regulations, 2011 CFR
2011-10-01
... form CG-706 or in the owner's format for an official logbook. Such logs must be kept available for a... master or person in charge shall file the logbook with the Officer in Charge, Marine Inspection. (b) The... of making entries therein as required by law or regulations in this subchapter. Such logs or records...
46 CFR 35.07-5 - Logbooks and records-TB/ALL.
Code of Federal Regulations, 2010 CFR
2010-10-01
..., the master or person in charge shall file the logbook with the Officer in Charge, Marine Inspection... purposes of making entries therein as required by law or regulations in this subchapter. Such logs or... records of tests and inspections of fire fighting equipment must be maintained with the vessel's logs for...
29 CFR 1960.28 - Employee reports of unsafe or unhealthful working conditions.
Code of Federal Regulations, 2011 CFR
2011-07-01
... report of an existing or potential unsafe or unhealthful working condition should be recorded on a log maintained at the establishment. If an agency finds it inappropriate to maintain a log of written reports at... sequentially numbered case file, coded for identification, should be assigned for purposes of maintaining an...
20 CFR 658.422 - Handling of non-JS-related complaints by the Regional Administrator.
Code of Federal Regulations, 2011 CFR
2011-04-01
... non-JS-related complaints alleging violations of employment related laws shall be logged. The... which the complainant (or complaint) was referred on a complaint log, similar to the one described in § 658.410(c)(1). The appropriate regional official shall also prepare and keep the file specified in...
46 CFR 35.07-5 - Logbooks and records-TB/ALL.
Code of Federal Regulations, 2011 CFR
2011-10-01
..., the master or person in charge shall file the logbook with the Officer in Charge, Marine Inspection... purposes of making entries therein as required by law or regulations in this subchapter. Such logs or... records of tests and inspections of fire fighting equipment must be maintained with the vessel's logs for...
NASA Technical Reports Server (NTRS)
Havelund, Klaus; Smith, Margaret H.; Barringer, Howard; Groce, Alex
2012-01-01
LogScope is a software package for analyzing log files. The intended use is for offline post-processing of such logs, after the execution of the system under test. LogScope can, however, in principle, also be used to monitor systems online during their execution. Logs are checked against requirements formulated as monitors expressed in a rule-based specification language. This language has similarities to a state machine language, but is more expressive, for example, in its handling of data parameters. The specification language is user friendly, simple, and yet expressive enough for many practical scenarios. The LogScope software was initially developed to specifically assist in testing JPL s Mars Science Laboratory (MSL) flight software, but it is very generic in nature and can be applied to any application that produces some form of logging information (which almost any software does).
Eldib, Ahmed; Li, Jinsheng; Emam, Ismail; Fan, Jiajin; Wang, Lu; Ma, C‐M
2014-01-01
The dosimetric advantage of modulated electron radiotherapy (MERT) has been explored by many investigators and is considered to be an advanced radiation therapy technique in the utilization of electrons. A computer‐controlled electron multileaf collimator (MLC) prototype, newly designed to be added onto a Varian linac to deliver MERT, was investigated both experimentally and by Monte Carlo simulations. Four different electron energies, 6, 9, 12, and 15 MeV, were employed for this investigation. To ensure that this device was capable of delivering the electron beams properly, measurements were performed to examine the electron MLC (eMLC) leaf leakage and to determine the appropriate jaw positioning for an eMLC‐shaped field in order to eliminate a secondary radiation peak that could otherwise appear outside of an intended radiation field in the case of inappropriate jaw positioning due to insufficient radiation blockage from the jaws. Phase space data were obtained by Monte Carlo (MC) simulation and recorded at the plane just above the jaws for each of the energies (6, 9, 12, and 15 MeV). As an input source, phase space data were used in MC dose calculations for various sizes of the eMLC shaped field (10×10 cm2, 3.4×3.4 cm2, and 2×2 cm2) with respect to a water phantom at source‐to‐surface distance (SSD)=94cm, while the jaws, eMLC leaves, and some accessories associated with the eMLC assembly as well were modeled as modifiers in the calculations. The calculated results were then compared with measurements from a water scanning system. The results showed that jaw settings with 5 mm margins beyond the field shaped by the eMLC were appropriate to eliminate the secondary radiation peak while not widening the beam penumbra; the eMLC leaf leakage measurements ranged from 0.3% to 1.8% for different energies based on in‐phantom measurements, which should be quite acceptable for MERT. Comparisons between MC dose calculations and measurements showed agreement within 1%/1mm based on percentage depth doses (PDDs) and off‐axis dose profiles for a range of field sizes for each of the electron energies. Our current work has demonstrated that the eMLC and other relevant components in the linac were correctly modeled and simulated via our in‐house MC codes, and the eMLC is capable of accurately delivering electron beams for various eMLC‐shaped field sizes with appropriate jaw settings. In the next stage, patient‐specific verification with a full MERT plan should be performed. PACS number: 87.55.ne PMID:24423848
33 CFR 1.07-90 - Criminal penalties.
Code of Federal Regulations, 2012 CFR
2012-07-01
... Commandant is required, the Area, Maintenance & Logistics Command (MLC), and District Commanders are..., 121 to 126 inclusive). (c) The Area, MLC, or District Commander will identify the laws or regulations...
33 CFR 1.07-90 - Criminal penalties.
Code of Federal Regulations, 2011 CFR
2011-07-01
... Commandant is required, the Area, Maintenance & Logistics Command (MLC), and District Commanders are..., 121 to 126 inclusive). (c) The Area, MLC, or District Commander will identify the laws or regulations...
33 CFR 1.07-90 - Criminal penalties.
Code of Federal Regulations, 2014 CFR
2014-07-01
... Commandant is required, the Area, Maintenance & Logistics Command (MLC), and District Commanders are..., 121 to 126 inclusive). (c) The Area, MLC, or District Commander will identify the laws or regulations...
33 CFR 1.07-90 - Criminal penalties.
Code of Federal Regulations, 2010 CFR
2010-07-01
... Commandant is required, the Area, Maintenance & Logistics Command (MLC), and District Commanders are..., 121 to 126 inclusive). (c) The Area, MLC, or District Commander will identify the laws or regulations...
33 CFR 1.07-90 - Criminal penalties.
Code of Federal Regulations, 2013 CFR
2013-07-01
... Commandant is required, the Area, Maintenance & Logistics Command (MLC), and District Commanders are..., 121 to 126 inclusive). (c) The Area, MLC, or District Commander will identify the laws or regulations...
Micro-Laser Range Finder Development: Using the Monolithic Approach
1999-02-01
components can be joined together, optically aligned to form the laser cavity and then sliced to produce MLC modules . This batch process can greatly reduce...the overall fabrication costs of the µLRF system. The MLC module is ultra-compact. Its overall size is approximately 56 mm (L) x 3 mm (W) x 3 mm (H) as...MLC module is placed on a laser pallet for stiffness, mechanical stability. The laser pallet size is selected as part of the integration design
McMahon, Ryan; Papiez, Lech; Rangaraj, Dharanipathy
2007-08-01
An algorithm is presented that allows for the control of multileaf collimation (MLC) leaves based entirely on real-time calculations of the intensity delivered over the target. The algorithm is capable of efficiently correcting generalized delivery errors without requiring the interruption of delivery (self-correcting trajectories), where a generalized delivery error represents anything that causes a discrepancy between the delivered and intended intensity profiles. The intensity actually delivered over the target is continually compared to its intended value. For each pair of leaves, these comparisons are used to guide the control of the following leaf and keep this discrepancy below a user-specified value. To demonstrate the basic principles of the algorithm, results of corrected delivery are shown for a leading leaf positional error during dynamic-MLC (DMLC) IMRT delivery over a rigid moving target. It is then shown that, with slight modifications, the algorithm can be used to track moving targets in real time. The primary results of this article indicate that the algorithm is capable of accurately delivering DMLC IMRT over a rigid moving target whose motion is (1) completely unknown prior to delivery and (2) not faster than the maximum MLC leaf velocity over extended periods of time. These capabilities are demonstrated for clinically derived intensity profiles and actual tumor motion data, including situations when the target moves in some instances faster than the maximum admissible MLC leaf velocity. The results show that using the algorithm while calculating the delivered intensity every 50 ms will provide a good level of accuracy when delivering IMRT over a rigid moving target translating along the direction of MLC leaf travel. When the maximum velocities of the MLC leaves and target were 4 and 4.2 cm/s, respectively, the resulting error in the two intensity profiles used was 0.1 +/- 3.1% and -0.5 +/- 2.8% relative to the maximum of the intensity profiles. For the same target motion, the error was shown to increase rapidly as (1) the maximum MLC leaf velocity was reduced below 75% of the maximum target velocity and (2) the system response time was increased.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Kraus, J; Thomas, E; Wu, X
2016-06-15
Purpose: Single-isocenter VMAT has been shown able to create high quality plans for complex intracranial multiple metastasis SRS cases. Linacs capable of the technique are typically outfitted with an MLC that consists of a combination of 5 mm and 10 mm leaves (standard) or 2.5 mm and 5 mm leaves (high-definition). In this study, we test the hypothesis that thinner collimator leaves are associated with improved plan quality. Methods: Ten multiple metastasis cases were identified and planned for VMAT SRS using a 10 MV flattening filter free beam. Plans were created for a standard (std) and a high-definition (HD) MLC.more » Published values for leaf transmission factor and dosimetric leaf gap were utilized. All other parameters were invariant. Conformity (plan and individual target), moderate isodose spill (V50%), and low isodose spill (mean brain dose) were selected for analysis. Results: Compared to standard MLC, HD-MLC improved overall plan conformity (median: Paddick CI-HD = 0.83, Paddick CI-std = 0.79; p = 0.004 and median: RTOG CI-HD =1.18, RTOG CI-std =1.24; p = 0.01 ), improved individual lesion conformity (median: Paddick CI-HD,i =0.77, Paddick CI-std,i =0.72; p < 0.001 and median: RTOG CI-HD,i = 1.28, RTOG CI-std,i =1.35; p < 0.001), improved moderate isodose spill (median: V50%-HD = 37.0 cc, V50%-std = 45.7 cc; p = 0.002), and improved low dose spill (median: dmean-HD = 2.90 Gy, dmean-std = 3.19 Gy; p = 0.002). Conclusion: For the single-isocenter VMAT SRS of multiple metastasis plans examined, use of HD-MLC modestly improved conformity, moderate isodose, and low isodose spill compared to standard MLC. However, in all cases we were able to generate clinically acceptable plans with the standard MLC. More work is need to further quantify the difference in cases with higher numbers of small targets and to better understand any potential clinical significance. This research was supported in part by Varian Medical Systems.« less
SU-D-201-04: Evaluation of Elekta Agility MLC Performance Using Statistical Process Control
DOE Office of Scientific and Technical Information (OSTI.GOV)
Meyers, SM; Balderson, MJ; Letourneau, D
2016-06-15
Purpose: to evaluate the performance and stability of the Elekta Agility MLC model using an automated quality control (QC) test in combination with statistical process control tools. Methods: Leaf positions were collected daily for 11 Elekta units over 5–19 months using the automated QC test, which analyzes 23 MV images to determine the location of MLC leaves relative to the radiation isocenter. The leaf positions are measured at 5 nominal positions, and images are acquired at collimator 0° and 180° to capture all MLC leaves in the field-of-view. Leaf positioning accuracy was assessed using individual and moving range control charts.more » Control limits were recomputed following MLC recalibration (occurred 1–2 times for 4 units). Specification levels of ±0.5, ±1 and ±1.5mm were tested. The mean and range of duration between out-of-control and out-of-specification events were determined. Results: Leaf position varied little over time, as confirmed by very tight individual control limits (mean ±0.19mm, range 0.09–0.44). Mean leaf position error was −0.03mm (range −0.89–0.83). Due to sporadic out-of-control events, the mean in-control duration was 3.3 days (range 1–23). Data stayed within ±1mm specification for 205 days on average (range 3–372) and within ±1.5mm for the entire date range. Measurements stayed within ±0.5mm for 1 day on average (range 0–17); however, our MLC leaves were not calibrated to this level of accuracy. Conclusion: The Elekta Agility MLC model was found to perform with high stability, as evidenced by the tight control limits. The in-specification durations support the current recommendation of monthly MLC QC tests with a ±1mm tolerance. Future work is on-going to determine if Agility performance can be optimized further using high-frequency QC test results to drive recalibration frequency. Factors that can affect leaf positioning accuracy, including beam spot motion, leaf gain calibration, drifting leaves, and image artifacts, are under investigation.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Li, X; Studenski, M; Yang, F
Purpose: MRI-guided-radiotherapy (MRIGRT) systems lack many features of traditional Linac based RT systems and specialized tests need to be developed to evaluate MLC performance. This work describes automatic tools for the analysis of positional accuracy of an MLC equipped MRIGRT system. Methods: This MLC analysis tool was developed for the MRIdian™ RT system which has three Co-60 equipped treatment heads each with a double focused MLC containing 30 leaf pairs, leaf thickness is 1.05cm defined at isocenter (SAD 105 cm). For MLC positional analysis a picket fence test was performed using a 25.4cm × 25.4cm Gafchromic™ RTQA2 film placed betweenmore » 5cm solidwater and a 30cm × 30cm × 1cm jigwire phantom with seven embedded parallel metal strips 4cm apart. A plan was generated to deliver 2Gy per field and seven 23.1cm × 2cm fields centered over each wire in the phantom. For each leaf pair the center of the radiation profile was determined by fitting the horizontal profile with a Gaussian model and determining the center of the FWHM. This was compared with the metal strip location to determine any deviation. The following metrics were used to evaluate the deviations per gantry angle including maximum, minimum, mean, Kurtosis, and skewness. Results: The identified maximum/mean leaf deviations are, 1.32/0.55 mm for gantry 0°, 1.59/0.76 mm for gantry 90°, and 1.19/0.39 mm for gantry 270°. The percentage of leaf deviation less than 1mm are 90.0% at 0°, 74.6% at 90°, and 97.0% at 270°. Kurtosis/skewness of the leaf deviation are 2.41/0.14 at 0°, 2.53/0.23 at 90°, 3.33/0.83 at 270°, respectively. Conclusion: This work presents an automatic tool for evaluation of the MLC position accuracy of the MRIdian™ radiotherapy system which can be used to benchmark the performance of the MLC system for each treatment head and track the results longitudinally.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Fürweger, Christoph, E-mail: christoph.fuerweger@cyber-knife.net; Prins, Paulette; Coskan, Harun
Purpose: The “InCise™ multileaf-collimator (MLC)” is the first commercial MLC to be mounted on a robotic SRS/SBRT platform (CyberKnife). The authors assessed characteristics and performance of this novel device in a preclinical five months test period. Methods: Commissioning beam data were acquired with unshielded diodes. EBT3 radiochromic films were employed for measurement of transmission, leaf/bank position accuracy (garden fence) before and after exercising the MLC, for end-to-end testing and further characterization of the beam. The robot workspace with MLC was assessed analytically by transformation to an Euler geometry (“plane,” “gantry,” and “collimator” angles) and by measuring pointing accuracy at eachmore » node. Stability over time was evaluated in picket fence and adapted Winston–Lutz tests (AQA). Results: Beam penumbrae (80%–20%, with 100% = 2 × dose at inflection point for field sizes ≥ 50 × 50 mm{sup 2}) were 2.2–3.7 mm for square fields in reference condition (source-axis-distance 800 mm, depth 15 mm) and depended on field size and off-axis position. Transmission and leakage did not exceed 0.5%. Accessible clinical workspace with MLC covered non-coplanar gantry angles of [−113°; +112°] and collimator angles of [−100°; +107°], with an average robot pointing accuracy of 0.12 ± 0.09 mm. For vertical beams, garden fence tests exhibited an average leaf positioning error of ≤0.2 mm, which increased by 0.25 and 0.30 mm (banks X1 and X2) with leaves traveling parallel to gravity. After execution of a leaf motion stress routine, garden fence tests showed slightly increased jaggedness and allowed to identify one malfunctioning leaf motor. Total system accuracy with MLC was 0.38 ± 0.05 mm in nine end-to-end tests. Picket fence and AQA tests displayed stable results over the test period. Conclusions: The InCise™ MLC for CyberKnife showed high accuracy and adequate characteristics for SRS/SBRT applications. MLC performance after exercise demands specific quality assurance measures.« less
Fürweger, Christoph; Prins, Paulette; Coskan, Harun; Heijmen, Ben J M
2016-05-01
The "InCise™ multileaf-collimator (MLC)" is the first commercial MLC to be mounted on a robotic SRS/SBRT platform (CyberKnife). The authors assessed characteristics and performance of this novel device in a preclinical five months test period. Commissioning beam data were acquired with unshielded diodes. EBT3 radiochromic films were employed for measurement of transmission, leaf/bank position accuracy (garden fence) before and after exercising the MLC, for end-to-end testing and further characterization of the beam. The robot workspace with MLC was assessed analytically by transformation to an Euler geometry ("plane," "gantry," and "collimator" angles) and by measuring pointing accuracy at each node. Stability over time was evaluated in picket fence and adapted Winston-Lutz tests (AQA). Beam penumbrae (80%-20%, with 100% = 2 × dose at inflection point for field sizes ≥ 50 × 50 mm(2)) were 2.2-3.7 mm for square fields in reference condition (source-axis-distance 800 mm, depth 15 mm) and depended on field size and off-axis position. Transmission and leakage did not exceed 0.5%. Accessible clinical workspace with MLC covered non-coplanar gantry angles of [-113°; +112°] and collimator angles of [-100°; +107°], with an average robot pointing accuracy of 0.12 ± 0.09 mm. For vertical beams, garden fence tests exhibited an average leaf positioning error of ≤0.2 mm, which increased by 0.25 and 0.30 mm (banks X1 and X2) with leaves traveling parallel to gravity. After execution of a leaf motion stress routine, garden fence tests showed slightly increased jaggedness and allowed to identify one malfunctioning leaf motor. Total system accuracy with MLC was 0.38 ± 0.05 mm in nine end-to-end tests. Picket fence and AQA tests displayed stable results over the test period. The InCise™ MLC for CyberKnife showed high accuracy and adequate characteristics for SRS/SBRT applications. MLC performance after exercise demands specific quality assurance measures.
2001-11-01
that there were· no· target misses. The Hellfire missile does not have a depleted uranium head . . -,, 2.2.2.3 Tank movement During the test, the...guide otber users through the use of this. complicated program. The_input data files for NOISEMAP consist of a root file name with several extensions...SOURCES subdirectory. This file will have the root file name followed by an accession number, then the .bps extension. The user must check the *.log
Tsukamoto, Takafumi; Yasunaga, Takuo
2014-11-01
Eos (Extensible object-oriented system) is one of the powerful applications for image processing of electron micrographs. In usual cases, Eos works with only character user interfaces (CUI) under the operating systems (OS) such as OS-X or Linux, not user-friendly. Thus, users of Eos need to be expert at image processing of electron micrographs, and have a little knowledge of computer science, as well. However, all the persons who require Eos does not an expert for CUI. Thus we extended Eos to a web system independent of OS with graphical user interfaces (GUI) by integrating web browser.Advantage to use web browser is not only to extend Eos with GUI, but also extend Eos to work under distributed computational environment. Using Ajax (Asynchronous JavaScript and XML) technology, we implemented more comfortable user-interface on web browser. Eos has more than 400 commands related to image processing for electron microscopy, and the usage of each command is different from each other. Since the beginning of development, Eos has managed their user-interface by using the interface definition file of "OptionControlFile" written in CSV (Comma-Separated Value) format, i.e., Each command has "OptionControlFile", which notes information for interface and its usage generation. Developed GUI system called "Zephyr" (Zone for Easy Processing of HYpermedia Resources) also accessed "OptionControlFIle" and produced a web user-interface automatically, because its mechanism is mature and convenient,The basic actions of client side system was implemented properly and can supply auto-generation of web-form, which has functions of execution, image preview, file-uploading to a web server. Thus the system can execute Eos commands with unique options for each commands, and process image analysis. There remain problems of image file format for visualization and workspace for analysis: The image file format information is useful to check whether the input/output file is correct and we also need to provide common workspace for analysis because the client is physically separated from a server. We solved the file format problem by extension of rules of OptionControlFile of Eos. Furthermore, to solve workspace problems, we have developed two type of system. The first system is to use only local environments. The user runs a web server provided by Eos, access to a web client through a web browser, and manipulate the local files with GUI on the web browser. The second system is employing PIONE (Process-rule for Input/Output Negotiation Environment), which is our developing platform that works under heterogenic distributed environment. The users can put their resources, such as microscopic images, text files and so on, into the server-side environment supported by PIONE, and so experts can write PIONE rule definition, which defines a workflow of image processing. PIONE run each image processing on suitable computers, following the defined rule. PIONE has the ability of interactive manipulation, and user is able to try a command with various setting values. In this situation, we contribute to auto-generation of GUI for a PIONE workflow.As advanced functions, we have developed a module to log user actions. The logs include information such as setting values in image processing, procedure of commands and so on. If we use the logs effectively, we can get a lot of advantages. For example, when an expert may discover some know-how of image processing, other users can also share logs including his know-hows and so we may obtain recommendation workflow of image analysis, if we analyze logs. To implement social platform of image processing for electron microscopists, we have developed system infrastructure, as well. © The Author 2014. Published by Oxford University Press on behalf of The Japanese Society of Microscopy. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.
Pitfalls of tungsten multileaf collimator in proton beam therapy.
Moskvin, Vadim; Cheng, Chee-Wai; Das, Indra J
2011-12-01
Particle beam therapy is associated with significant startup and operational cost. Multileaf collimator (MLC) provides an attractive option to improve the efficiency and reduce the treatment cost. A direct transfer of the MLC technology from external beam radiation therapy is intuitively straightforward to proton therapy. However, activation, neutron production, and the associated secondary cancer risk in proton beam should be an important consideration which is evaluated. Monte Carlo simulation with FLUKA particle transport code was applied in this study for a number of treatment models. The authors have performed a detailed study of the neutron generation, ambient dose equivalent [H∗(10)], and activation of a typical tungsten MLC and compared with those obtained from a brass aperture used in a typical proton therapy system. Brass aperture and tungsten MLC were modeled by absorber blocks in this study, representing worst-case scenario of a fully closed collimator. With a tungsten MLC, the secondary neutron dose to the patient is at least 1.5 times higher than that from a brass aperture. The H∗(10) from a tungsten MLC at 10 cm downstream is about 22.3 mSv/Gy delivered to water phantom by noncollimated 200 MeV beam of 20 cm diameter compared to 14 mSv/Gy for the brass aperture. For a 30-fraction treatment course, the activity per unit volume in brass aperture reaches 5.3 × 10⁴ Bq cm(-3) at the end of the last treatment. The activity in brass decreases by a factor of 380 after 24 h, additional 6.2 times after 40 days of cooling, and is reduced to background level after 1 yr. Initial activity in tungsten after 30 days of treating 30 patients per day is about 3.4 times higher than in brass that decreases only by a factor of 2 after 40 days and accumulates to 1.2 × 10⁶ Bq cm(-3) after a full year of operation. The daily utilization of the MLC leads to buildup of activity with time. The overall activity continues to increase due to (179)Ta with a half-life of 1.82 yr and thus require prolonged storage for activity cooling. The H∗(10) near the patient side of the tungsten block is about 100 μSv/h and is 27 times higher at the upstream side of the block. This would lead to an accumulated dose for therapists in a year that may exceed occupational maximum permissible dose (50 mSv/yr). The value of H∗(10) at the upstream surface of the tungsten block is about 220 times higher than that of the brass. MLC is an efficient way for beam shaping and overall cost reduction device in proton therapy. However, based on this study, tungsten seems to be not an optimal material for MLC in proton beam therapy. Usage of tungsten MLC in clinic may create unnecessary risks associated with the secondary neutrons and induced radioactivity for patients and staff depending on the patient load. A careful selection of material for manufacturing of an optimal MLC for proton therapy is thus desired.
25 CFR 215.23 - Cooperation between superintendent and district mining supervisor.
Code of Federal Regulations, 2011 CFR
2011-04-01
... notices, reports, drill logs, maps, and records, and all other information relating to mining operations required by said regulations to be submitted by lessees, and shall maintain a file thereof for the superintendent. (b) The files of the Geological Survey supervisor relating to lead and zinc leases of Quapaw...
Agentless Cloud-Wide Monitoring of Virtual Disk State
2015-10-01
packages include Apache, MySQL , PHP, Ruby on Rails, Java Application Servers, and many others. Figure 2.12 shows the results of a run of the Software...Linux, Apache, MySQL , PHP (LAMP) set of applications. Thus, many file-level update logs will contain the same versions of files repeated across many
Military Standard Common APSE (Ada Programming Support Environment) Interface Set (CAIS).
1985-01-01
QUEUEASE. LAST-KEY (QUEENAME) . LASTREI.TIONI(QUEUE-NAME). FILE-NODE. PORN . ATTRIBUTTES. ACCESSCONTROL. LEVEL); CLOSE (QUEUE BASE); CLOSE(FILE NODE...PROPOSED XIIT-STD-C.4 31 J NNUAfY logs procedure zTERT (ITERATOR: out NODE ITERATON; MAMIE: NAME STRING.KIND: NODE KID : KEY : RELATIONSHIP KEY PA1TTE1 :R
Rius-Vilarrasa, E; Bünger, L; Maltin, C; Matthews, K R; Roehe, R
2009-05-01
The Meat and Livestock Commission's (MLC) EUROP classification based scheme and Video Image Analysis (VIA) system were compared in their ability to predict weights of primal carcass joints. A total of 443 commercial lamb carcasses under 12 months of age and mixed gender were selected by their cold carcass weight (CCW), conformation and fat scores. Lamb carcasses were classified for conformation and fatness, scanned by the VIA system and dissected into primal joints of leg, chump, loin, breast and shoulder. After adjustment for CCW, the estimation of primal joints using MLC EUROP scores showed high coefficients of determination (R(2)) in the range of 0.82-0.99. The use of VIA always resulted in equal or higher R(2). The precision measured as root mean square error (RMSE) was 27% (leg), 13% (chump), 1% (loin), 11% (breast), 5% (shoulders) and 13% (total primals) higher using VIA than MLC carcass information. Adjustment for slaughter day and gender effects indicated that estimations of primal joints using MLC EUROP scores were more sensitive to these factors than using VIA. This was consistent with an increase in stability of the prediction model of 28%, 11%, 2%, 12%, 6% and 14% for leg, chump, loin, breast and shoulder and total primals, respectively, using VIA compared to MLC EUROP scores. Consequently, VIA was capable of improving the prediction of primal meat yields compared to the current MLC EUROP carcass classification scheme used in the UK abattoirs.
47 CFR 76.1704 - Proof-of-performance test data.
Code of Federal Regulations, 2011 CFR
2011-10-01
...-performance test data. (a) The proof of performance tests required by § 76.601 shall be maintained on file at... subscribers, subject to the requirements of § 76.601(d). Note to § 76.1704: If a signal leakage log is being... log must be retained for the period specified in § 76.601(d). ...
49 CFR Appendix A to Part 225 - Schedule of Civil Penalties 1
Code of Federal Regulations, 2010 CFR
2010-10-01
... $1,000 $2,000 225.11Reports of accidents/ incidents 2,500 5,000 225.12(a): Failure to file Railroad... noncompliance: (1) a missing or incomplete log entry for a particular employee's injury or illness; or (2) a missing or incomplete log record for a particular rail equipment accident or incident. Each day a...
47 CFR 76.1704 - Proof-of-performance test data.
Code of Federal Regulations, 2010 CFR
2010-10-01
...-performance test data. (a) The proof of performance tests required by § 76.601 shall be maintained on file at... subscribers, subject to the requirements of § 76.601(d). Note to § 76.1704: If a signal leakage log is being... log must be retained for the period specified in § 76.601(d). ...
ERIC Educational Resources Information Center
Avouris, N.; Fiotakis, G.; Kahrimanis, G.; Margaritis, M.; Komis, V.
2007-01-01
In this article, we discuss key requirements for collecting behavioural data concerning technology-supported collaborative learning activities. It is argued that the common practice of analysis of computer generated log files of user interactions with software tools is not enough for building a thorough view of the activity. Instead, more…
A Clustering Methodology of Web Log Data for Learning Management Systems
ERIC Educational Resources Information Center
Valsamidis, Stavros; Kontogiannis, Sotirios; Kazanidis, Ioannis; Theodosiou, Theodosios; Karakos, Alexandros
2012-01-01
Learning Management Systems (LMS) collect large amounts of data. Data mining techniques can be applied to analyse their web data log files. The instructors may use this data for assessing and measuring their courses. In this respect, we have proposed a methodology for analysing LMS courses and students' activity. This methodology uses a Markov…
Consistency of Students' Pace in Online Learning
ERIC Educational Resources Information Center
Hershkovitz, Arnon; Nachmias, Rafi
2009-01-01
The purpose of this study is to investigate the consistency of students' behavior regarding their pace of actions over sessions within an online course. Pace in a session is defined as the number of logged actions divided by session length (in minutes). Log files of 6,112 students were collected, and datasets were constructed for examining pace…
Diffenderfer, Eric S; Ainsley, Christopher G; Kirk, Maura L; McDonough, James E; Maughan, Richard L
2011-11-01
To apply the dual ionization chamber method for mixed radiation fields to an accurate comparison of the secondary neutron dose arising from the use of a tungsten alloy multileaf collimator (MLC) as opposed to a brass collimator system for defining the shape of a therapeutic proton field. Hydrogenous and nonhydrogenous ionization chambers were constructed with large volumes to enable measurements of absorbed doses below 10(-4) Gy in mixed radiation fields using the dual ionization chamber method for mixed-field dosimetry. Neutron dose measurements were made with a nominal 230 MeV proton beam incident on a closed tungsten alloy MLC and a solid brass block. The chambers were cross-calibrated against a (60)Co-calibrated Farmer chamber in water using a 6 MV x-ray beam and Monte Carlo simulations were performed to account for variations in ionization chamber response due to differences in secondary neutron energy spectra. The neutron and combined proton plus γ-ray absorbed doses are shown to be nearly equivalent downstream from either a closed tungsten alloy MLC or a solid brass block. At 10 cm downstream from the distal edge of the collimating material the neutron dose from the closed MLC was (5.3 ± 0.4) × 10(- 5) Gy/Gy. The neutron dose with brass was (6.4 ± 0.7) × 10(- 5) Gy/Gy. Further from the secondary neutron source, at 50 cm, the neutron doses remain close for both the MLC and brass block at (6.9 ± 0.6) × 10(- 6) Gy/Gy and (6.3 ± 0.7) × 10(- 6) Gy/Gy, respectively. The dual ionization chamber method is suitable for measuring secondary neutron doses resulting from proton irradiation. The results of measurements downstream from a closed tungsten alloy MLC and a brass block indicate that, even in an overly pessimistic worst-case scenario, secondary neutron production in a tungsten alloy MLC leads to absorbed doses that are nearly equivalent to those seen from brass collimators. Therefore, the choice of tungsten alloy in constructing the leaves of a proton MLC is appropriate, and does not lead to a substantial increase in the secondary neutron dose to the patient compared to that generated in a brass collimator.
Satpathy, M; Gallagher, P; Lizotte-Waniewski, M; Srinivas, S P
2004-10-01
Phosphorylation of the regulatory light chain of myosin II (referred to as myosin light chain or MLC) leads to a loss of barrier integrity in cellular monolayers by an increase in the contractility of the cortical actin cytoskeleton. This effect has been examined in corneal endothelial (CE) cells. Experiments were performed using cultured bovine CE cells (BCEC). MLC phosphorylation was induced by a thrombin-mediated activation of the proteinase-activated receptor-1 (PAR-1). Expression of MLC kinase (MLCK), a Ca2+/calmodulin-dependent protein kinase that phosphorylates MLC at its Ser-19 and Thr-18 residues, was determined by RT-PCR and Western blotting. Expression of PAR-1, RhoA, and Rho kinase-1 (effector of RhoA) was ascertained by RT-PCR. MLC phosphorylation was assessed by urea-glycerol gel electrophoresis followed by immunoblotting. The effects of Rho kinase-1 and PKC were characterized by using their selective inhibitors, Y-27632 and chelerythrine, respectively. Reorganization of the cytoskeleton was evaluated by the phalloidin staining of actin. [Ca2+]i was measured using Fura-2. The barrier integrity was assayed as permeability of BCEC monolayers to horseradish peroxidase (HRP; 44 kDa). RT-PCR showed expression of MLCK, PAR-1, Rho kinase-1, and RhoA. Western blotting indicated expression of the non-muscle and smooth muscle isoforms of MLCK. Exposure to thrombin induced an increase in [Ca2+]i with the peak unaffected by an absence of extracellular Ca2+. Pre-exposure to thrombin (2 U ml(-1); 2 min) led to mono- and di-phosphorylation of MLC. Under both basal conditions and in the presence of thrombin, MLC phosphorylation was prevented by chelerythrine (10 microm) and Y-27632 (<25 microm). Thrombin led to inter-endothelial gaps secondary to the disruption of the cortical actin cytoskeleton, which under resting conditions was organized as a perijunctional actomyosin ring (PAMR). These responses were blocked by pre-treatment with Y-27632. Thrombin also increased permeability to HRP, which was abolished by pre-treatment with Y-27632. Thrombin induces MLC phosphorylation in BCEC. The consequent increase in the contractility of the actin cytoskeleton produces a centripetal force resulting in inter-endothelial gaps and a breakdown of barrier integrity. These responses are PKC- and Rho kinase-dependent. [Ca2+]i increase, as well as sensitivity of the thrombin response to PKC and Rho kinase inhibitors, are consistent with the expression of PAR-1 receptors in BCEC. Thrombin-induced hyperpermeability is a model to investigate barrier dysfunction induced by MLC phosphorylation.
Raper, J E
1977-01-01
Since February 1976, The Medical Library Center of New York, with the assistance of the SUNY/OCLC Network, has offered, on a subscription basis, a centralized automated cataloging service to health science libraries in the greater metropolitan New York area. By using workforms and prints of OCLC record (amended by the subscribing participants), technical services personnel at the center have fed cataloging data, via a CRT terminal, into the OCLC system, which provides (1) catalog cards, received in computer filing order; (2) book card, spine, and pocket labels; (3) accessions lists; and (4) data for eventual production of book catalogs and union catalogs. The experience of the center in the development, implementation, operation, and budgeting of its shared cataloging service is discussed. PMID:843650
Raper, J E
1977-04-01
Since February 1976, The Medical Library Center of New York, with the assistance of the SUNY/OCLC Network, has offered, on a subscription basis, a centralized automated cataloging service to health science libraries in the greater metropolitan New York area. By using workforms and prints of OCLC record (amended by the subscribing participants), technical services personnel at the center have fed cataloging data, via a CRT terminal, into the OCLC system, which provides (1) catalog cards, received in computer filing order; (2) book card, spine, and pocket labels; (3) accessions lists; and (4) data for eventual production of book catalogs and union catalogs. The experience of the center in the development, implementation, operation, and budgeting of its shared cataloging service is discussed.
Greer, Peter B; Vial, Philip; Oliver, Lyn; Baldock, Clive
2007-11-01
The aim of this work was to experimentally determine the difference in response of an amorphous silicon (a-Si) electronic portal imaging device (EPID) to the open and multileaf collimator (MLC) transmitted beam components of intensity modulated radiation therapy (IMRT) beams. EPID dose response curves were measured for open and MLC transmitted (MLCtr) 10 x 10 cm2 beams at central axis and with off axis distance using a shifting field technique. The EPID signal was obtained by replacing the flood-field correction with a pixel sensitivity variation matrix correction. This signal, which includes energy-dependent response, was then compared to ion-chamber measurements. An EPID calibration method to remove the effect of beam energy variations on EPID response was developed for IMRT beams. This method uses the component of open and MLCtr fluence to an EPID pixel calculated from the MLC delivery file and applies separate radially dependent calibration factors for each component. The calibration procedure does not correct for scatter differences between ion chamber in water measurements and EPID response; these must be accounted for separately with a kernel-based approach or similar method. The EPID response at central axis for the open beam was found to be 1.28 +/- 0.03 of the response for the MLCtr beam, with the ratio increasing to 1.39 at 12.5 cm off axis. The EPID response to MLCtr radiation did not change with off-axis distance. Filtering the beam with copper plates to reduce the beam energy difference between open and MLCtr beams was investigated; however, these were not effective at reducing EPID response differences. The change in EPID response for uniform sliding window IMRT beams with MLCtr dose components from 0.3% to 69% was predicted to within 2.3% using the separate EPID response calibration factors for each dose component. A clinical IMRT image calibrated with this method differed by nearly 30% in high MLCtr regions from an image calibrated with an open beam calibration factor only. Accounting for the difference in EPID response to open and MLCtr radiation should improve IMRT dosimetry with a-Si EPIDs.
Analysis of the Impact of Data Normalization on Cyber Event Correlation Query Performance
2012-03-01
2003). Organizations use it in planning, target marketing , decision-making, data analysis, and customer services (Shin, 2003). Organizations that...Following this IP address is a router message sequence number. This is a globally unique number for each router terminal and can range from...Appendix G, invokes the PERL parser for the log files from a particular USAF base, and invokes the CTL file that loads the resultant CSV file into the
Sawmill: A Logging File System for a High-Performance RAID Disk Array
1995-01-01
from limiting disk performance, new controller architectures connect the disks directly to the network so that data movement bypasses the file server...These developments raise two questions for file systems: how to get the best performance from a RAID, and how to use such a controller architecture ...the RAID-II storage system; this architecture provides a fast data path that moves data rapidly among the disks, high-speed controller memory, and the
2014-10-07
is counted as. Per the TDTC, a test bridge with longitudinal and/or lateral symmetry under non- eccentric loading can be considered as 1, 2, or 4...Level Run036 3 MLC70T (tracked) BA Run046 6 AB Run055 9 AB Run060 9 BA Run064 12 BA Run071 15 AB Run155 3 MLC96W ( wheeled ) AB...Run331 9 AB Run359 15 AB Run430 12 MLC96W ( wheeled ) BA Run434 12 AB Run447 3 BA Bank Condition: Side Slope, Even Strain Channels High
NASA Astrophysics Data System (ADS)
Yoon, Do-Kun; Jung, Joo-Young; Suh, Tae Suk
2014-05-01
In order to confirm the possibility of field application of a different type collimator with a multileaf collimator (MLC), we constructed a grid-type multi-layer pixel collimator (GTPC) by using a Monte Carlo n-particle simulation (MCNPX). In this research, a number of factors related to the performance of the GPTC were evaluated using simulated output data of a basic MLC model. A layer was comprised of a 1024-pixel collimator (5.0 × 5.0 mm2) which could operate individually as a grid-type collimator (32 × 32). A 30-layer collimator was constructed for a specific portal form to pass radiation through the opening and closing of each pixel cover. The radiation attenuation level and the leakage were compared between the GTPC modality simulation and MLC modeling (tungsten, 17.50 g/cm3, 5.0 × 70.0 × 160.0 mm3) currently used for a radiation field. Comparisons of the portal imaging, the lateral dose profile from a virtual water phantom, the dependence of the performance on the increase in the number of layers, the radiation intensity modulation verification, and the geometric error between the GTPC and the MLC were done using the MCNPX simulation data. From the simulation data, the intensity modulation of the GTPC showed a faster response than the MLC's (29.6%). In addition, the agreement between the doses that should be delivered to the target region was measured as 97.0%, and the GTPC system had an error below 0.01%, which is identical to that of MLC. A Monte Carlo simulation of the GTPC could be useful for verification of application possibilities. Because the line artifact is caused by the grid frame and the folded cover, a lineal dose transfer type is chosen for the operation of this system. However, the result of GTPC's performance showed that the methods of effective intensity modulation and the specific geometric beam shaping differed with the MLC modality.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Lee, Young; Ruschin, Mark
SRS using linac and cones offers steep dose fall-off but a tradeoff exists between conformality and treatment time, which depends on the number of isocentres. Purpose of this study is to quantify planning metrics between cones- and MLC-based SRS for arteriovenous malformation(AVM). Seven AVM cases treated with cones were re-planned with MLC on Pinnacle treatment planning system. Planning target volume(PTV) was created with 1mm uniform margin to the AVM to account for MLC positional variation. Clinically-planned prescription dose(15–25Gy) was used. Four plans were generated per case:non-coplanar VMAT(ncV), single-arc VMAT(saV), non-coplanar IMRT(ncI), non-coplanar conformal(ncC). Plans were compared for conformity(CI), heterogeneity(HI) andmore » gradient(GI) indices and brain doses. Estimated treatment times and monitor units(MU) were compared. Cone-based plans required 2–6 isocentres. Though CI-RTOG was similar for plans(median=0.98), CI-Paddick was most favourable for ncV(median=0.86) and worst for cones(0.54). HI for MLC plans(median=1.19–1.27) were lower than cone-based plans(1.43). GI was similar for all plans. For 2/7 ncC had brainstem maximum dose>16.7Gy and therefore were clinically unacceptable. Brain V12Gy,V10Gy,V2Gy were lowest in the cones plan. ncV brain V12Gy,V10Gy,V2Gy were lowest of all MLC-based plans studied. Treatment MUs were similar for MLC-based plans and up to 70% lower than clinically delivered plans. ncV showed best conformality in this study. Of the MLC-based plans, ncV also showed lowest normal tissue dose with reasonable treatment time.« less
Kwon, Ji-Wook; Kim, Jin Hyo; Seo, Jiwon
2015-01-01
This paper proposes a Multiple Leader Candidate (MLC) structure and a Competitive Position Allocation (CPA) algorithm which can be applicable for various applications including environmental sensing. Unlike previous formation structures such as virtual-leader and actual-leader structures with position allocation including a rigid allocation and an optimization based allocation, the formation employing the proposed MLC structure and CPA algorithm is robust against the fault (or disappearance) of the member robots and reduces the entire cost. In the MLC structure, a leader of the entire system is chosen among leader candidate robots. The CPA algorithm is the decentralized position allocation algorithm that assigns the robots to the vertex of the formation via the competition of the adjacent robots. The numerical simulations and experimental results are included to show the feasibility and the performance of the multiple robot system employing the proposed MLC structure and the CPA algorithm. PMID:25954956
32 CFR 776.80 - Initial screening and Rules Counsel.
Code of Federal Regulations, 2011 CFR
2011-07-01
... Director, JA Division, HQMC, to JAR. (b) JAG(13) and JAR shall log all complaints received and will ensure... within 30 days of the date of its return, the Rules Counsel may close the file without further action... action to close the file. (2) Complaints that comply with the requirements shall be further reviewed by...
Coastal bathymetry data collected in 2011 from the Chandeleur Islands, Louisiana
DeWitt, Nancy T.; Pfeiffer, William R.; Bernier, Julie C.; Buster, Noreen A.; Miselis, Jennifer L.; Flocks, James G.; Reynolds, Billy J.; Wiese, Dana S.; Kelso, Kyle W.
2014-01-01
This report serves as an archive of processed interferometric swath and single-beam bathymetry data. Geographic Iinformation System data products include a 50-meter cell-size interpolated bathymetry grid surface, trackline maps, and point data files. Additional files include error analysis maps, Field Activity Collection System logs, and formal Federal Geographic Data Committee metadata.
Online Courses Assessment through Measuring and Archetyping of Usage Data
ERIC Educational Resources Information Center
Kazanidis, Ioannis; Theodosiou, Theodosios; Petasakis, Ioannis; Valsamidis, Stavros
2016-01-01
Database files and additional log files of Learning Management Systems (LMSs) contain an enormous volume of data which usually remain unexploited. A new methodology is proposed in order to analyse these data both on the level of both the courses and the learners. Specifically, "regression analysis" is proposed as a first step in the…
Impact of the MLC on the MRI field distortion of a prototype MRI-linac
DOE Office of Scientific and Technical Information (OSTI.GOV)
Kolling, Stefan; Keall, Paul; Oborn, Brad
2013-12-15
Purpose: To cope with intrafraction tumor motion, integrated MRI-linac systems for real-time image guidance are currently under development. The multileaf collimator (MLC) is a key component in every state-of-the-art radiotherapy treatment system, allowing for accurate field shaping and tumor tracking. This work quantifies the magnetic impact of a widely used MLC on the MRI field homogeneity for such a modality.Methods: The finite element method was employed to model a MRI-linac assembly comprised of a 1.0 T split-bore MRI magnet and the key ferromagnetic components of a Varian Millennium 120 MLC, namely, the leaves and motors. Full 3D magnetic field maps ofmore » the system were generated. From these field maps, the peak-to-peak distortion within the MRI imaging volume was evaluated over a 30 cm diameter sphere volume (DSV) around the isocenter and compared to a maximum preshim inhomogeneity of 300 μT. Five parametric studies were performed: (1) The source-to-isocenter distance (SID) was varied from 100 to 200 cm, to span the range of a compact system to that with lower magnetic coupling. (2) The MLC model was changed from leaves only to leaves with motors, to determine the contribution to the total distortion caused by MLC leaves and motors separately. (3) The system was configured in the inline or perpendicular orientation, i.e., the linac treatment beam was oriented parallel or perpendicular to the magnetic field direction. (4) The treatment field size was varied from 0 × 0 to 20×20 cm{sup 2}, to span the range of clinical treatment fields. (5) The coil currents were scaled linearly to produce magnetic field strengths B{sub 0} of 0.5, 1.0, and 1.5 T, to estimate how the MLC impact changes with B{sub 0}.Results: (1) The MLC-induced MRI field distortion fell continuously with increasing SID. (2) MLC leaves and motors were found to contribute to the distortion in approximately equal measure. (3) Due to faster falloff of the fringe field, the field distortion was generally smaller in the perpendicular beam orientation. The peak-to-peak DSV distortion was below 300 μT at SID≥130 cm (perpendicular) and SID≥140 cm (inline) for the 1.0 T design. (4) The simulation of different treatment fields was identified to cause dynamic changes in the field distribution. However, the estimated residual distortion was below 1.2 mm geometric distortion at SID≥120 cm (perpendicular) and SID≥130 cm (inline) for a 10 mT/m frequency-encoding gradient. (5) Due to magnetic saturation of the MLC materials, the field distortion remained constant at B{sub 0}>1.0 T.Conclusions: This work shows that the MRI field distortions caused by the MLC cannot be ignored and must be thoroughly investigated for any MRI-linac system. The numeric distortion values obtained for our 1.0 T magnet may vary for other magnet designs with substantially different fringe fields, however the concept of modest increases in the SID to reduce the distortion to a shimmable level is generally applicable.« less
Predicting Correctness of Problem Solving from Low-Level Log Data in Intelligent Tutoring Systems
ERIC Educational Resources Information Center
Cetintas, Suleyman; Si, Luo; Xin, Yan Ping; Hord, Casey
2009-01-01
This paper proposes a learning based method that can automatically determine how likely a student is to give a correct answer to a problem in an intelligent tutoring system. Only log files that record students' actions with the system are used to train the model, therefore the modeling process doesn't require expert knowledge for identifying…
Code of Federal Regulations, 2011 CFR
2011-01-01
..., Import Inspection Division, is on file at the import inspection facility where the inspection is to be... stamping log containing the following information for each lot of product: the date of inspection, the... container marks, and the MP-410 number covering the product to be inspected. The daily stamping log must be...
ERIC Educational Resources Information Center
Gobert, Janice D.; Sao Pedro, Michael; Raziuddin, Juelaila; Baker, Ryan S.
2013-01-01
We present a method for assessing science inquiry performance, specifically for the inquiry skill of designing and conducting experiments, using educational data mining on students' log data from online microworlds in the Inq-ITS system (Inquiry Intelligent Tutoring System; www.inq-its.org). In our approach, we use a 2-step process: First we use…
Genetics Home Reference: megalencephalic leukoencephalopathy with subcortical cysts
... is unable to correctly transport GlialCAM and MLC1 proteins to cell junctions. It is unknown how a lack of functional MLC1 or GlialCAM protein at cell junctions in the brain impairs brain development and ...
Quantitative measurement of MLC leaf displacements using an electronic portal image device
NASA Astrophysics Data System (ADS)
Yang, Yong; Xing, Lei
2004-04-01
The success of an IMRT treatment relies on the positioning accuracy of the MLC (multileaf collimator) leaves for both step-and-shoot and dynamic deliveries. In practice, however, there exists no effective and quantitative means for routine MLC QA and this has become one of the bottleneck problems in IMRT implementation. In this work we present an electronic portal image device (EPID) based method for fast and accurate measurement of MLC leaf positions at arbitrary locations within the 40 cm × 40 cm radiation field. The new technique utilizes the fact that the integral signal in a small region of interest (ROI) is a sensitive and reliable indicator of the leaf displacement. In this approach, the integral signal at a ROI was expressed as a weighted sum of the contributions from the displacements of the leaf above the point and the adjacent leaves. The weighting factors or linear coefficients of the system equations were determined by fitting the integral signal data for a group of pre-designed MLC leaf sequences to the known leaf displacements that were intentionally introduced during the creation of the leaf sequences. Once the calibration is done, the system can be used for routine MLC leaf positioning QA to detect possible leaf errors. A series of tests was carried out to examine the functionality and accuracy of the technique. Our results show that the proposed technique is potentially superior to the conventional edge-detecting approach in two aspects: (i) it deals with the problem in a systematic approach and allows us to take into account the influence of the adjacent MLC leaves effectively; and (ii) it may improve the signal-to-noise ratio and is thus capable of quantitatively measuring extremely small leaf positional displacements. Our results indicate that the technique can detect a leaf positional error as small as 0.1 mm at an arbitrary point within the field in the absence of EPID set-up error and 0.3 mm when the uncertainty is considered. Given its simplicity, efficiency and accuracy, we believe that the technique is ideally suitable for routine MLC leaf positioning QA. This work was presented at the 45th Annual Meeting of American Society of Therapeutic Radiology and Oncology (ASTRO), Salt Lake City, UT, 2003. A US Patent is pending (application no. 10/197,232).
Determination of MLC model parameters for Monaco using commercial diode arrays.
Kinsella, Paul; Shields, Laura; McCavana, Patrick; McClean, Brendan; Langan, Brian
2016-07-08
Multileaf collimators (MLCs) need to be characterized accurately in treatment planning systems to facilitate accurate intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT). The aim of this study was to examine the use of MapCHECK 2 and ArcCHECK diode arrays for optimizing MLC parameters in Monaco X-ray voxel Monte Carlo (XVMC) dose calculation algorithm. A series of radiation test beams designed to evaluate MLC model parameters were delivered to MapCHECK 2, ArcCHECK, and EBT3 Gafchromic film for comparison. Initial comparison of the calculated and ArcCHECK-measured dose distributions revealed it was unclear how to change the MLC parameters to gain agreement. This ambiguity arose due to an insufficient sampling of the test field dose distributions and unexpected discrepancies in the open parts of some test fields. Consequently, the XVMC MLC parameters were optimized based on MapCHECK 2 measurements. Gafchromic EBT3 film was used to verify the accuracy of MapCHECK 2 measured dose distributions. It was found that adjustment of the MLC parameters from their default values resulted in improved global gamma analysis pass rates for MapCHECK 2 measurements versus calculated dose. The lowest pass rate of any MLC-modulated test beam improved from 68.5% to 93.5% with 3% and 2 mm gamma criteria. Given the close agreement of the optimized model to both MapCHECK 2 and film, the optimized model was used as a benchmark to highlight the relatively large discrepancies in some of the test field dose distributions found with ArcCHECK. Comparison between the optimized model-calculated dose and ArcCHECK-measured dose resulted in global gamma pass rates which ranged from 70.0%-97.9% for gamma criteria of 3% and 2 mm. The simple square fields yielded high pass rates. The lower gamma pass rates were attributed to the ArcCHECK overestimating the dose in-field for the rectangular test fields whose long axis was parallel to the long axis of the ArcCHECK. Considering ArcCHECK measurement issues and the lower gamma pass rates for the MLC-modulated test beams, it was concluded that MapCHECK 2 was a more suitable detector than ArcCHECK for the optimization process. © 2016 The Authors
DOE Office of Scientific and Technical Information (OSTI.GOV)
Wang, L; Chin, E; Lo, A
2016-06-15
Purpose: This work reports the results of the physics evaluation of a newly released InCise™2 Multileaf Collimator (MLC) installed in our institution. Methods: Beam property data was measured with unshielded diode and EBT2 films. The measurements included MLC leaf transmission, beam profiles, output factors and tissue-phantom ratios. MLC performance was evaluated for one month after commissioning. Weekly Garden Fence tests were performed for leaf / bank positioning in standard (A/P) and clinically relevant non-standard positions, before and after MLC driving exercises of 10+ minutes. Daily Picket Fence test and AQA test, End-to-End tests and dosimetric quality assurance were performed tomore » evaluate the overall system performance. Results: All measurements including beam energy, flatness and symmetry, were within manufacture specifications. Leaf transmission was 0.4% <0.5% specification. The values of output factors ranged from 0.825 (7.6 mm × 7.5 mm) to 1.026 (115.0 mm × 100.1 mm). Average beam penumbra at 10 cm depth ranged from 2.7mm/2.7mm(7.6 mm × 7.5 mm) to 6.0 mm/6.2mm(84.6 mm × 84.7 mm). Slight penumbra difference (<10% from average penumbra for fields >20 mm) was observed in the direction perpendicular to leaf motion due to the tilting of the leaf housing. Mean leaf position offsets was −0.08±0.07mm and −0.13 ± 0.08 for X1 and X2 leaf banks in 13 Garden Fence tests. No significant difference on average leaf positioning offsets was observed between different leaf orientations and before/after MLC driving exercises. Six End-to-End tests showed 0.43±0.23mm overall targeting accuracy. Picket-Fence and AQA showed stable performance of MLC during the test period. Dosimetric point dose measurements for test cases agreed with calculation within 3%. All film measurements on relative dose had Gamma (2%, 2mm) passing rate of >95%. Conclusion: The Incise™2 MLC for CyberKnife M6™ was proven to be accurate and reliable, and it is currently in clinical use. Stanford was one of the physics evaluation sites for the newly released InCise 2 MLC for Accuray Inc.« less
Visual behavior characterization for intrusion and misuse detection
NASA Astrophysics Data System (ADS)
Erbacher, Robert F.; Frincke, Deborah
2001-05-01
As computer and network intrusions become more and more of a concern, the need for better capabilities, to assist in the detection and analysis of intrusions also increase. System administrators typically rely on log files to analyze usage and detect misuse. However, as a consequence of the amount of data collected by each machine, multiplied by the tens or hundreds of machines under the system administrator's auspices, the entirety of the data available is neither collected nor analyzed. This is compounded by the need to analyze network traffic data as well. We propose a methodology for analyzing network and computer log information visually based on the analysis of the behavior of the users. Each user's behavior is the key to determining their intent and overriding activity, whether they attempt to hide their actions or not. Proficient hackers will attempt to hide their ultimate activities, which hinders the reliability of log file analysis. Visually analyzing the users''s behavior however, is much more adaptable and difficult to counteract.
Parallel file system with metadata distributed across partitioned key-value store c
Bent, John M.; Faibish, Sorin; Grider, Gary; Torres, Aaron
2017-09-19
Improved techniques are provided for storing metadata associated with a plurality of sub-files associated with a single shared file in a parallel file system. The shared file is generated by a plurality of applications executing on a plurality of compute nodes. A compute node implements a Parallel Log Structured File System (PLFS) library to store at least one portion of the shared file generated by an application executing on the compute node and metadata for the at least one portion of the shared file on one or more object storage servers. The compute node is also configured to implement a partitioned data store for storing a partition of the metadata for the shared file, wherein the partitioned data store communicates with partitioned data stores on other compute nodes using a message passing interface. The partitioned data store can be implemented, for example, using Multidimensional Data Hashing Indexing Middleware (MDHIM).
Improving cardiovascular health of underserved populations in the community with Life's Simple 7.
Murphy, Marcia Pencak; Coke, Lola; Staffileno, Beth A; Robinson, Janis D; Tillotson, Robin
2015-11-01
The purpose of this nurse practitioner (NP) led initiative was to improve the cardiovascular health of two underserved populations in the community using the American Heart Association (AHA) Life's Simple 7 and My Life Check (MLC) tools. Two inner city community sites were targeted: (a) a senior center servicing African American (AA) older adults, and (b) a residential facility servicing homeless women. Preprogram health data (blood pressure, cholesterol, blood glucose levels, body mass index, and health behaviors) were collected to calculate MLC scores. Postprogram health data were obtained on participants with the lowest MLC scores who completed the program. Eight older adults completed the program with a 37.1% increase in average MLC score (6.2 vs. 8.5). Ten women completed the program with a 9.3% decrease in average MLC score (4.3 vs. 3.9). Favorable benefits were observed in the AA older adults. In contrast, similar benefits were not observed in the women, which may be because of a constellation of social, environmental, biological, and mental health factors. NPs are prepared to target community-based settings to address the health of underserved populations. Engaging key stakeholders in the planning and implementation is essential for success. ©2015 American Association of Nurse Practitioners.
The effect of iron on metronidazole activity against Trichomonas vaginalis in vitro.
Elwakil, Hala Salah; Tawfik, Rania Ayman; Alam-Eldin, Yosra Hussein; Nassar, Doaa Ashraf
2017-11-01
Metronidazole is administered in an inactive form then activated to its cytotoxic form within the hydrogenosome of trichomonads. Two hydrogenosomal proteins, pyruvate ferredoxin oxidoreductase (PFOR) and ferredoxin, play a critical role in the reductive activation of metronidazole. The expression of these proteins and other hydrogenosomal proteins are likewise positively regulated by iron. In the present study, the effect of iron on minimal lethal concentration (MLC) of metronidazole on in vitro cultured Trichomonas vaginalis(T. vaginalis) isolates was investigated. Interestingly, Addition of Ferrous ammonium sulphate (FAS) to T. vaginalis culture led to decrease in the MLC of metronidazole. On using aerobic assay, MLC of metronidazole on untreated T. vaginalis of both isolates was 12.5 μg/ml that decreased to 0.38 μg/ml on FAS treated trichomonads. Also anaerobic assay revealed that MLC on untreated parasites was 3.12 μg/ml that decreased to 0.097 μg/ml and 0.19 μg/ml for isolate 1 and isolate 2 respectively after iron addition. It was concluded that, addition of iron to in vitro cultured T. vaginalis decreases metronidazole MLC that was detected by both aerobic and anaerobic assays. Copyright © 2017 Elsevier Inc. All rights reserved.
Vercelli, D; Jabara, H H; Lauener, R P; Geha, R S
1990-01-15
The T cell-derived lymphokine IL-4 is essential for the induction of IgE synthesis by human lymphocytes. The IgE-inducing effect of IL-4 is antagonized by IFN-gamma. The secretion of IFN-gamma is vigorously triggered in MLC. Thus, IL-4-stimulated MLC represent a suitable model to characterize the functional antagonism between IL-4 and IFN-gamma. In this report, we show that rIL-4 consistently induced IgE synthesis when added to human primary MLC. IL-4-dependent IgE production required cognate T/B cell recognition, because it was inhibited by antibodies to CD3 and MHC class II (HlA-DR) Ag. A neutralizing anti-IFN-gamma mAb dramatically enhanced IL-4-dependent IgE synthesis by MLC, indicating that endogenous IFN-gamma is a major inhibitor of IgE production. More importantly, addition of rIL-4 markedly inhibited the release of IFN-gamma in supernatants of MLC and Con A-activated PBMC. The decrease in IFN-gamma protein was accompanied by a decreased expression of IFN-gamma mRNA transcripts. The downregulation of IFN-gamma by IL-4 is likely to play an important role in the IL-4-dependent induction of IgE synthesis.
An increase or a decrease in myosin II phosphorylation inhibits macrophage motility
1991-01-01
Myosin II purified from mammalian non-muscle cells is phosphorylated on the 20-kD light chain subunit (MLC20) by the Ca2+/calmodulin-dependent enzyme myosin light chain kinase (MLCK). The importance of MLC20 phosphorylation in regulating cell motility was investigated by introducing either antibodies to MLCK (MK-Ab) or a Ca2+/calmodulin- independent, constitutively active form of MLCK (MK-) into macrophages. The effects of these proteins on cell motility were then determined using a quantitative chemotaxis assay. Chemotaxis is significantly diminished in macrophages containing MK-Ab compared to macrophages containing control antibodies. Moreover, there is an inverse relationship between the number of cells that migrate and the amount of MK-Ab introduced into cells. Interestingly, there is also an inverse relationship between the number of cells that migrate and the amount of MK- introduced into cells. Other experiments demonstrated that MK-Ab decreased intracellular MLC20 phosphorylation while MK- increased MLC20 phosphorylation. MK- also increased the amount of myosin associated with the cytoskeleton. These data demonstrate that the regulation of MLCK is an important aspect of cell motility and suggest that MLC20 phosphorylation must be maintained within narrow limits during translational motility by mammalian cells. PMID:2071674
Georgiou, Stella; Alami-Durante, Hélène; Power, Deborah M; Sarropoulou, Elena; Mamuris, Zissis; Moutou, Katerina A
2016-02-01
Hyperplasia and hypertrophy are the two mechanisms by which muscle develops and grows. We study these two mechanisms, during the early development of white muscle in Sparus aurata, by means of histology and the expression of structural and regulatory genes. A clear stage of stratified hyperplasia was identified early in the development of gilthead sea bream but ceased by 35 dph when hypertrophy took over. Mosaic recruitment of new white fibers began as soon as 60 dph. The genes mlc2a and mlc2b were expressed at various levels during the main phases of hyperplasia and hypertrophy. The genes myog and mlc2a were significantly up-regulated during the intensive stratified formation of new fibers and their expression was significantly correlated. Expression of mstn1 and igf1 increased at 35 dph, appeared to regulate the hyperplasia-to-hypertrophy transition, and may have stimulated the expression of mlc2a, mlc2b and col1a1 at the onset of mosaic hyperplasia. The up-regulation of mstn1 at transitional phases in muscle development indicates a dual regulatory role of myostatin in fish larval muscle growth.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Fuerweger, C; European Cyberknife Center Munich, Munich, DE; Prins, P
Purpose: To assess characteristics and performance of the “Incise™” MLC (41 leaf pairs, 2.5mm width, FFF linac) mounted on the robotic SRS/SBRT platform “CyberKnife M6™” in a pre-clinical 5 months (11/2014–03/2015) test period. Methods: Beam properties were measured with unshielded diodes and EBT3 film. The CyberKnife workspace for MLC was analyzed by transforming robot node coordinates (cranial / body paths) into Euler geometry. Bayouth tests for leaf / bank position accuracy were performed in standard (A/P) and clinically relevant non-standard positions, before and after exercising the MLC for 10+ minutes. Total system and delivery accuracy were assessed in End-to-End testsmore » and dosimetric verification of exemplary plans. Stability over time was evaluated in Picket-Fence-and adapted Winston-Lutz-tests (AQA) for different collimator angles. Results: Penumbrae (80–20%, with 100%=2*dose at inflection point; SAD 80cm; 10cm depth) parallel / perpendicular to leaf motion were 2.87/2.64mm for the smallest (0×76×0.75cm{sup 2}) and 5.34/4.94mm for the largest (9.76×9.75cm{sup 2}) square field. MLC circular field penumbrae exceeded fixed cones by 10–20% (e.g. 60mm: 4.0 vs. 3.6mm; 20mm: 3.6 vs. 2.9mm). Interleaf leakage was <0.5%. Clinically accessible workspace with MLC covered (non-coplanar) gantry angles of [-113°;+112°] (cranial) and [-108°;+102°] (body), and collimator angles of [-100°;+107°] (cranial) and [-91°;+100°] (body). Average leaf position offsets were ≤0.2mm in 14 standard A/P Bayouth tests and ≤0.6mm in 8 non-standard direction tests. Pre-test MLC exercise increased jaggedness (range ±0.3mm vs. ±0.5mm) and allowed to identify one malfunctioning leaf motor. Total system accuracy with MLC was 0.39±0.06mm in 6 End-to-End tests. Picket-Fence and AQA showed no adverse trends during the test period. Conclusion: The Incise™ MLC for CyberKnife M6™ displayed high accuracy and mechanical stability over the test period. The specific CyberKnife geometry and performance after exercise demand dedicated QA measures. This work is in part funded by a research grant from Accuray Inc, Sunnyvale, USA. Erasmus MC Cancer Institute also has research collaborations with Elekta AB, Stockholm, Sweden. C Fuerweger has previously received speaker honoraria from Accuray Inc, Sunnyvale, USA.« less
Ground-water data for the Hanna and Carbon basins, south-central Wyoming, through 1980
Daddow, P.B.
1986-01-01
Groundwater resources in the Hanna and Carbon Basins of Wyoming were assessed in a study from 1974 through 1980 because of the development of coal mining in the area. Data collected from 105 wells during that study, including well-completion records, lithologic logs, and water levels, are presented. The data are from stock wells, coal-test holes completed as observation wells by the U.S. Geological Survey. The data are mostly from mined coal-bearing formations: the Tertiary Hanna Formation and the Tertiary and Cretaceous Ferris Formation. Well-completion data and lithologic logs were collected on-site during drilling of the wells or from U.S. Geological Survey files, company records, Wyoming State Engineer well-permit files, and published reports. (USGS)
VizieR Online Data Catalog: The Gemini Observation Log (CADC, 2001-)
NASA Astrophysics Data System (ADS)
Association of Universities For Research in Astronomy
2018-01-01
This database contains a log of the Gemini Telescope observations since 2001, managed by the Canadian Astronomical Data Center (CADC). The data are regularly updated (see the date of the last version at the end of this file). The Gemini Observatory consists of twin 8.1-meter diameter optical/infrared telescopes located on two of the best observing sites on the planet. From their locations on mountains in Hawai'i and Chile, Gemini Observatory's telescopes can collectively access the entire sky. Gemini is operated by a partnership of five countries including the United States, Canada, Brazil, Argentina and Chile. Any astronomer in these countries can apply for time on Gemini, which is allocated in proportion to each partner's financial stake. (1 data file).
REPHLEX II: An information management system for the ARS Water Data Base
NASA Astrophysics Data System (ADS)
Thurman, Jane L.
1993-08-01
The REPHLEX II computer system is an on-line information management system which allows scientists, engineers, and other researchers to retrieve data from the ARS Water Data Base using asynchronous communications. The system features two phone lines handling baud rates from 300 to 2400, customized menus to facilitate browsing, help screens, direct access to information and data files, electronic mail processing, file transfers using the XMODEM protocol, and log-in procedures which capture information on new users, process passwords, and log activity for a permanent audit trail. The primary data base on the REPHLEX II system is the ARS Water Data Base which consists of rainfall and runoff data from experimental agricultural watersheds located in the United States.
Ontology based log content extraction engine for a posteriori security control.
Azkia, Hanieh; Cuppens-Boulahia, Nora; Cuppens, Frédéric; Coatrieux, Gouenou
2012-01-01
In a posteriori access control, users are accountable for actions they performed and must provide evidence, when required by some legal authorities for instance, to prove that these actions were legitimate. Generally, log files contain the needed data to achieve this goal. This logged data can be recorded in several formats; we consider here IHE-ATNA (Integrating the healthcare enterprise-Audit Trail and Node Authentication) as log format. The difficulty lies in extracting useful information regardless of the log format. A posteriori access control frameworks often include a log filtering engine that provides this extraction function. In this paper we define and enforce this function by building an IHE-ATNA based ontology model, which we query using SPARQL, and show how the a posteriori security controls are made effective and easier based on this function.
A Prototype Implementation of a Time Interval File Protection System in Linux
2006-09-01
when a user logs in, the /etc/ passwd file is read by the system to get the user’s home directory. The user’s login shell then changes the directory...and don. • Users can be added with the command: # useradd – m <username> • Set the password by: # passwd <username> • Make a copy of the
Well 9-1 Logs and Data: Roosevelt Hot Spring Area, Utah (FORGE)
Joe Moore
2016-03-03
This is a compilation of logs and data from Well 9-1 in the Roosevelt Hot Springs area in Utah. This well is also in the Utah FORGE study area. The file is in a compressed .zip format and there is a data inventory table (Excel spreadsheet) in the root folder that is a guide to the data that is accessible in subfolders.
ERIC Educational Resources Information Center
Asunka, Stephen; Chae, Hui Soo; Hughes, Brian; Natriello, Gary
2009-01-01
Transaction logs of user activity on an academic library website were analyzed to determine general usage patterns on the website. This paper reports on insights gained from the analysis, and identifies and discusses issues relating to content access, interface design and general functionality of the website. (Contains 13 figures and 8 tables.)
The Use of OPAC in a Large Academic Library: A Transactional Log Analysis Study of Subject Searching
ERIC Educational Resources Information Center
Villen-Rueda, Luis; Senso, Jose A.; de Moya-Anegon, Felix
2007-01-01
The analysis of user searches in catalogs has been the topic of research for over four decades, involving numerous studies and diverse methodologies. The present study looks at how different types of users effect queries in the catalog of a university library. For this purpose, we analyzed log files to determine which was the most frequent type of…
ERIC Educational Resources Information Center
Cho, Moon-Heum; Yoo, Jin Soung
2017-01-01
Many researchers who are interested in studying students' online self-regulated learning (SRL) have heavily relied on self-reported surveys. Data mining is an alternative technique that can be used to discover students' SRL patterns from large data logs saved on a course management system. The purpose of this study was to identify students' online…
ERIC Educational Resources Information Center
Copeland, Tom
Figuring depreciation can be the most difficult aspect of filing tax returns for a family child care program. This inventory log for family child care programs is designed to assist in keeping track of the furniture, appliances, and other property used in the child care business; once these items have been identified, they can be deducted as…
DOE Office of Scientific and Technical Information (OSTI.GOV)
Carosi, Alessandra, E-mail: alessandra.carosi@katamail.com; Ingrosso, Gianluca; Ponti, Elisabetta
2014-07-01
The purpose of this study is to analyze the dosimetric effect of Elekta Beam Modulator in 3-dimensional conformal radiation therapy (3DCRT) and in intensity-modulated radiation therapy (IMRT) for localized prostate cancer. We compared treatment plans developed with 2 different Elekta multileaf collimators (MLC): Beam Modulator micro-MLC (mMLC) (4-mm leaf width at the isocenter) and standard MLC (10-mm leaf width at the isocenter). The comparison was performed for 15 patients with localized prostate cancer in 3DCRT and IMRT delivery; a total of 60 treatment plans were processed. The dose-volume histograms were used to provide the quantitative comparison between plans. In particular,more » we analyzed differences between rectum and bladder sparing in terms of a set of appropriate Vx (percentage of organ at risk [OAR] volume receiving the x dose) and differences between target conformity and coverage in terms of coverage factor and conformation number. Our analysis demonstrates that in 3DCRT there is an advantage in the use of Elekta Beam Modulator mMLC in terms of organ sparing; in particular, a significant decrease in rectal V{sub 60} and V{sub 50} (p = 0.001) and in bladder V{sub 70} and V{sub 65} (p = 0.007 and 0.002, respectively) was found. Moreover, a better target dose conformity was obtained (p = 0.002). IMRT plans comparison demonstrated no significant differences between the use of the 4 or 10-mm MLCs. Our analysis shows that in 3DCRT the use of the Elekta Beam Modulator mMLC gives a gain in target conformity and in OARs dose sparing whereas in IMRT plans there is no advantage.« less
Myosin light chain phosphorylation is critical for adaptation to cardiac stress.
Warren, Sonisha A; Briggs, Laura E; Zeng, Huadong; Chuang, Joyce; Chang, Eileen I; Terada, Ryota; Li, Moyi; Swanson, Maurice S; Lecker, Stewart H; Willis, Monte S; Spinale, Francis G; Maupin-Furlowe, Julie; McMullen, Julie R; Moss, Richard L; Kasahara, Hideko
2012-11-27
Cardiac hypertrophy is a common response to circulatory or neurohumoral stressors as a mechanism to augment contractility. When the heart is under sustained stress, the hypertrophic response can evolve into decompensated heart failure, although the mechanism(s) underlying this transition remain largely unknown. Because phosphorylation of cardiac myosin light chain 2 (MLC2v), bound to myosin at the head-rod junction, facilitates actin-myosin interactions and enhances contractility, we hypothesized that phosphorylation of MLC2v plays a role in the adaptation of the heart to stress. We previously identified an enzyme that predominantly phosphorylates MLC2v in cardiomyocytes, cardiac myosin light-chain kinase (cMLCK), yet the role(s) played by cMLCK in regulating cardiac function in health and disease remain to be determined. We found that pressure overload induced by transaortic constriction in wild-type mice reduced phosphorylated MLC2v levels by ≈40% and cMLCK levels by ≈85%. To examine how a reduction in cMLCK and the corresponding reduction in phosphorylated MLC2v affect function, we generated Mylk3 gene-targeted mice and transgenic mice overexpressing cMLCK specifically in cardiomyocytes. Pressure overload led to severe heart failure in cMLCK knockout mice but not in mice with cMLCK overexpression in which cMLCK protein synthesis exceeded degradation. The reduction in cMLCK protein during pressure overload was attenuated by inhibition of ubiquitin-proteasome protein degradation systems. Our results suggest the novel idea that accelerated cMLCK protein turnover by the ubiquitin-proteasome system underlies the transition from compensated hypertrophy to decompensated heart failure as a result of reduced phosphorylation of MLC2v.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Thiyagarajan, Rajesh; Karrthick, KP; Kataria, Tejinder
Purpose: Performing DQA for Bilateral (B-L) breast tomotherapy is a challenging task due to the limitation of any commercially available detector array or film. Aim of this study is to perform DQA for B-L breast tomotherapy plan using MLC fluence sinogram. Methods: Treatment plan was generated on Tomotherapy system for B-L breast tumour. B-L breast targets were given 50.4 Gy prescribed over 28 fractions. Plan is generated with 6 MV photon beam & pitch was set to 0.3. As the width of the total target is 39 cm (left & right) length is 20 cm. DQA plan delivered without anymore » phantom on the mega voltage computed tomography (MCVT) detector system. The pulses recorded by MVCT system were exported to the delivery analysis software (Tomotherapy Inc.) for reconstruction. The detector signals are reconstructed to a sonogram and converted to MLC fluence sonogram. The MLC fluence sinogram compared with the planned fluence sinogram. Also point dose measured with cheese phantom and ionization chamber to verify the absolute dose component Results: Planned fluence sinogram and reconstructed MLC fluence sinogram were compared using Gamma metric. MLC positional difference and intensity of the beamlet were used as parameters to evaluate gamma. 3 mm positional difference and 3% beamlet intensity difference were used set for gamma calculation. A total of 26784 non-zero beamlets were included in the analysis out of which 161 beamlets had gamma more than 1. The gamma passing rate found to be 99.4%. Point dose measurements were within 1.3% of the calculated dose. Conclusion: MLC fluence sinogram based delivery quality assurance performed for bilateral breast irradiation. This would be a suitable alternate for large volume targets like bilateral breast, Total body irradiation etc. However conventional method of DQA should be used to validate this method periodically.« less
Ait-Belgnaoui, A; Han, W; Lamine, F; Eutamene, H; Fioramonti, J; Bueno, L; Theodorou, V
2006-08-01
Stress induced increase in colonic paracellular permeability results from epithelial cell cytoskeleton contraction and is responsible for stress induced hypersensitivity to colorectal distension (CRD). The probiotic Lactobacillus farciminis releases spontaneously nitric oxide (NO) in the colonic lumen in vivo and exerts anti-inflammatory effects. This study aimed: (i) to evaluate the effects of L farciminis on stress induced hypersensitivity to CRD and increase in colonic paracellular permeability; and (ii) to ascertain whether these effects are NO mediated and related to changes in colonocyte myosin light chain phosphorylation (p-MLC). Female Wistar rats received either 10(11) CFU/day of L farciminis or saline orally over 15 days before partial restraint stress (PRS) or sham-PRS application. Visceral sensitivity to CRD and colonic paracellular permeability was assessed after PRS or sham-PRS. Haemoglobin was used as an NO scavenger. Western blotting for MLC kinase, MLC, and p-MLC were performed in colonic mucosa from L farciminis treated and control rats after PRS or sham-PRS. PRS significantly increased the number of spike bursts for CRD pressures of 30-60 mm Hg as well as colonic paracellular permeability. L farciminis treatment prevented both effects, while haemoglobin reversed the protective effects of L farciminis. p-MLC expression increased significantly from 15 to 45 minutes after PRS, and L farciminis treatment prevented this increase. L farciminis treatment prevents stress induced hypersensitivity, increase in colonic paracellular permeability, and colonocyte MLC phosphorylation. This antinociceptive effect occurs via inhibition of contraction of colonic epithelial cell cytoskeleton and the subsequent tight junction opening, and may also involve direct or indirect effects of NO produced by this probiotic.
Ait‐Belgnaoui, A; Han, W; Lamine, F; Eutamene, H; Fioramonti, J; Bueno, L; Theodorou, V
2006-01-01
Background Stress induced increase in colonic paracellular permeability results from epithelial cell cytoskeleton contraction and is responsible for stress induced hypersensitivity to colorectal distension (CRD). The probiotic Lactobacillus farciminis releases spontaneously nitric oxide (NO) in the colonic lumen in vivo and exerts anti‐inflammatory effects. This study aimed: (i) to evaluate the effects of L farciminis on stress induced hypersensitivity to CRD and increase in colonic paracellular permeability; and (ii) to ascertain whether these effects are NO mediated and related to changes in colonocyte myosin light chain phosphorylation (p‐MLC). Methods Female Wistar rats received either 1011 CFU/day of L farciminis or saline orally over 15 days before partial restraint stress (PRS) or sham‐PRS application. Visceral sensitivity to CRD and colonic paracellular permeability was assessed after PRS or sham‐PRS. Haemoglobin was used as an NO scavenger. Western blotting for MLC kinase, MLC, and p‐MLC were performed in colonic mucosa from L farciminis treated and control rats after PRS or sham‐PRS. Results PRS significantly increased the number of spike bursts for CRD pressures of 30–60 mm Hg as well as colonic paracellular permeability. L farciminis treatment prevented both effects, while haemoglobin reversed the protective effects of L farciminis. p‐MLC expression increased significantly from 15 to 45 minutes after PRS, and L farciminis treatment prevented this increase. Conclusion L farciminis treatment prevents stress induced hypersensitivity, increase in colonic paracellular permeability, and colonocyte MLC phosphorylation. This antinociceptive effect occurs via inhibition of contraction of colonic epithelial cell cytoskeleton and the subsequent tight junction opening, and may also involve direct or indirect effects of NO produced by this probiotic. PMID:16507583
Jackson, Nathan D.; Austin, Christopher C.
2013-01-01
Despite considerable attention, the long-term impact of rivers on species diversification remains uncertain. Meander loop cutoff (MLC) is one river phenomenon that may compromise a river’s diversifying effects by passively transferring organisms from one side of the river to the other. However, the ability of MLC to promote gene flow across rivers has not been demonstrated empirically. Here, we test several predictions of MLC-mediated gene flow in populations of North American ground skinks (Scincella lateralis) separated by a well-established riverine barrier, the Mississippi River: 1) individuals collected from within meander cutoffs should be more closely related to individuals across the river than on the same side, 2) individuals within meander cutoffs should contain more immigrants than individuals away from meander cutoffs, 3) immigration rates estimated across the river should be highest in the direction of the cutoff event, and 4) the distribution of alleles native to one side of the river should be better predicted by the historical rather than current path of the river. To test these predictions we sampled 13 microsatellite loci and mitochondrial DNA from ground skinks collected near three ancient meander loops. These predictions were generally supported by genetic data, although support was stronger for mtDNA than for microsatellite data. Partial support for genetic divergence of samples within ancient meander loops also provides evidence for the MLC hypothesis. Although a role for MLC-mediated gene flow was supported here for ground skinks, the transient nature of river channels and morphologies may limit the long-term importance of MLC in stemming population divergence across major rivers. PMID:23658778
NOTE: Reducing the number of segments in unidirectional MLC segmentations
NASA Astrophysics Data System (ADS)
Mellado, X.; Cruz, S.; Artacho, J. M.; Canellas, M.
2010-02-01
In intensity-modulated radiation therapy (IMRT), fluence matrices obtained from a treatment planning system are usually delivered by a linear accelerator equipped with a multileaf collimator (MLC). A segmentation method is needed for decomposing these fluence matrices into segments suitable for the MLC, and the number of segments used is an important factor for treatment time. In this work, an algorithm for reduction of the number of segments (NS) is presented for unidirectional segmentations, where there is no backtracking of the MLC leaves. It uses a geometrical representation of the segmentation output for searching the key values in a fluence matrix that complicate its decomposition. The NS reduction is achieved by performing minor modifications in these values, under the conditions of avoiding substantial modifications of the dose-volume histogram, and does not increase in average the total number of monitor units delivered. The proposed method was tested using two clinical cases planned with the PCRT 3D® treatment planning system.
Wister, CA Downhole and Seismic Data
Akerley, John
2010-12-18
This submission contains Downhole geophysical logs associated with Wister, CA Wells 12-27 and 85-20. The logs include Spontaneous Potential (SP), HILT Caliper (HCAL), Gamma Ray (GR), Array Induction (AIT), and Neutron Porosity (NPOR) data. Also included are a well log, Injection Test, Pressure Temperature Spinner log, shut in temperature survey, a final well schematic, and files about the well's location and drilling history. This submission also contains data from a three-dimensional (3D) multi-component (3C) seismic reflection survey on the Wister Geothermal prospect area in the northern portion of the Imperial Valley, California. The Wister seismic survey area was 13.2 square miles. (Resistivity image logs (Schlumberger FMI) in 85-20 indicate that maximum horizontal stress (Shmax) is oriented NNE but that open fractures are oriented suboptimally).
SU-E-T-01: 2-D Characterization of DLG Among All MLC Leaf Pairs
DOE Office of Scientific and Technical Information (OSTI.GOV)
Kumaraswamy, L; Xu, Z; Podgorsak, M
Purpose: The aim of this study is to evaluate the variation of dosimetric leaf-gap (DLG) along the travel path of each MLC leaf pair. This study evaluates whether the spatial variations in DLG could cause dose differences between TPS-calculated and measured dose. Methods: The 6MV DLG values were measured for all leaf pairs in the direction of leaf motion using a 2-D diode array and 0.6cc ion chamber. These measurements were performed on two Varian Linacs, employing the Millennium 120-leaf MLC and a 2-D-DLG variation map was created via in-house software. Several test plans were created with sweeping MLC fieldsmore » using constant gaps from 2mm to 10mm and corrected for 2-D variation utilizing in-house software. Measurements were performed utilizing the MapCHECK at 5.0cm depth for plans with and without the 2-D DLG correction and compared to the TPS calculated dose via gamma analysis (3%/3mm). Results: The measured DLGs for the middle 40 MLC leaf pairs (0.5cm width) were very similar along the central superior-inferior axis, with maximum variation of 0.2mm. The outer 20 MLC leaf pairs (1.0cm width) have DLG values from 0.32mm (mean) to 0.65mm (maximum) lower than the central leaf-pair, depending on off-axis distance. Gamma pass rates for the 2mm, 4mm, and 6mm sweep plans increased by 23.2%, 28.7%, and 26.0% respectively using the 2-D-DLG correction. The most improved dose points occur in areas modulated by the 1.0cm leaf-pairs. The gamma pass rate for the 10mm sweep plan increased by only 7.7%, indicating that the 2D variation becomes less significant for dynamic plans with larger MLC gaps. Conclusion: Fluences residing significantly off-axis with narrow sweeping gaps may exhibit significant variations from planned dose due to large differences between the true DLG exhibited by the 1.0cm leaf-pairs versus the constant DLG value utilized by the TPS for dose calculation.« less
Leung, Jacqueline M.; Tran, Fanny; Pathak, Ravindra B.; Poupart, Séverine; Heaslip, Aoife T.; Ballif, Bryan A.; Westwood, Nicholas J.; Ward, Gary E.
2014-01-01
Motility of the protozoan parasite Toxoplasma gondii plays an important role in the parasite’s life cycle and virulence within animal and human hosts. Motility is driven by a myosin motor complex that is highly conserved across the Phylum Apicomplexa. Two key components of this complex are the class XIV unconventional myosin, TgMyoA, and its associated light chain, TgMLC1. We previously showed that treatment of parasites with a small-molecule inhibitor of T. gondii invasion and motility, tachypleginA, induces an electrophoretic mobility shift of TgMLC1 that is associated with decreased myosin motor activity. However, the direct target(s) of tachypleginA and the molecular basis of the compound-induced TgMLC1 modification were unknown. We show here by “click” chemistry labelling that TgMLC1 is a direct and covalent target of an alkyne-derivatized analogue of tachypleginA. We also show that this analogue can covalently bind to model thiol substrates. The electrophoretic mobility shift induced by another structural analogue, tachypleginA-2, was associated with the formation of a 225.118 Da adduct on S57 and/or C58, and treatment with deuterated tachypleginA-2 confirmed that the adduct was derived from the compound itself. Recombinant TgMLC1 containing a C58S mutation (but not S57A) was refractory to click labelling and no longer exhibited a mobility shift in response to compound treatment, identifying C58 as the site of compound binding on TgMLC1. Finally, a knock-in parasite line expressing the C58S mutation showed decreased sensitivity to compound treatment in a quantitative 3D motility assay. These data strongly support a model in which tachypleginA and its analogues inhibit the motility of T. gondii by binding directly and covalently to C58 of TgMLC1, thereby causing a decrease in the activity of the parasite’s myosin motor. PMID:24892871
DOE Office of Scientific and Technical Information (OSTI.GOV)
Boehnke, E McKenzie; DeMarco, J; Steers, J
2016-06-15
Purpose: To examine both the IQM’s sensitivity and false positive rate to varying MLC errors. By balancing these two characteristics, an optimal tolerance value can be derived. Methods: An un-modified SBRT Liver IMRT plan containing 7 fields was randomly selected as a representative clinical case. The active MLC positions for all fields were perturbed randomly from a square distribution of varying width (±1mm to ±5mm). These unmodified and modified plans were measured multiple times each by the IQM (a large area ion chamber mounted to a TrueBeam linac head). Measurements were analyzed relative to the initial, unmodified measurement. IQM readingsmore » are analyzed as a function of control points. In order to examine sensitivity to errors along a field’s delivery, each measured field was divided into 5 groups of control points, and the maximum error in each group was recorded. Since the plans have known errors, we compared how well the IQM is able to differentiate between unmodified and error plans. ROC curves and logistic regression were used to analyze this, independent of thresholds. Results: A likelihood-ratio Chi-square test showed that the IQM could significantly predict whether a plan had MLC errors, with the exception of the beginning and ending control points. Upon further examination, we determined there was ramp-up occurring at the beginning of delivery. Once the linac AFC was tuned, the subsequent measurements (relative to a new baseline) showed significant (p <0.005) abilities to predict MLC errors. Using the area under the curve, we show the IQM’s ability to detect errors increases with increasing MLC error (Spearman’s Rho=0.8056, p<0.0001). The optimal IQM count thresholds from the ROC curves are ±3%, ±2%, and ±7% for the beginning, middle 3, and end segments, respectively. Conclusion: The IQM has proven to be able to detect not only MLC errors, but also differences in beam tuning (ramp-up). Partially supported by the Susan Scott Foundation.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Chibani, O; Tahanout, F; Ma, C
2016-06-15
Purpose: To commission a new MLC model for the GEPTS Monte Carlo system. The model is based on the concept of leaves and interleaves effective densities Methods: GEPTS is a Monte Carlo system to be used for external beam planning verification. GEPTS incorporates detailed photon and electron transport algorithms (Med.Phys. 29, 2002, 835). A new GEPTS model for the Varian Millennium MLC is presented. The model accounts for: 1) thick (1 cm) and thin (0.5 cm) leaves, 2) tongue-and-groove design, 3) High-Transmission (HT) and Low-Transmission (LT) interleaves, and 4) rounded leaf end. Leaf (and interleaf) height is set equal tomore » 6 cm. Instead of modeling air gaps, screw holes, and complex leaf heads, “effective densities” are assigned to: 1) thin leaves, 2) thick leaves, 3) HT-, and 4) LT-interleaves. Results: The new MLC model is used to calculate dose profiles for Closed-MLC and Tongue-and-Groove fields at 5 cm depth for 6, 10 and 15 MV Varian beams. Calculations are compared with 1) Pin-point ionization chamber transmission ratios and 2) EBT3 Radiochromic films. Pinpoint readings were acquired beneath thick and thin leaves, and HT and LT interleaves. The best fit of measured dose profiles was obtained for the following parameters: Thick-leaf density = 16.1 g/cc, Thin-leaf density = 17.2 g/cc; HT Interleaf density = 12.4 g/cc, LT Interleaf density = 14.3 g/cc; Interleaf thickness = 1.1 mm. Attached figures show comparison of calculated and measured transmission ratios for the 3 energies. Note this is the only study where transmission profiles are compared with measurements for 3 different energies. Conclusion: The new MLC model reproduces transmission measurements within 0.1%. The next step is to implement the MLC model for real plans and quantify the improvement in dose calculation accuracy gained using this model for IMRT plans with high modulation factors.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Pogson, E; Liverpool and Macarthur Cancer Therapy Centres, Liverpool, NSW; Ingham Institute for Applied Medical Research, Sydney, NSW
Purpose: To quantify the impact of differing magnitudes of simulated linear accelerator errors on the dose to the target volume and organs at risk for nasopharynx VMAT. Methods: Ten nasopharynx cancer patients were retrospectively replanned twice with one full arc VMAT by two institutions. Treatment uncertainties (gantry angle and collimator in degrees, MLC field size and MLC shifts in mm) were introduced into these plans at increments of 5,2,1,−1,−2 and −5. This was completed using an in-house Python script within Pinnacle3 and analysed using 3DVH and MatLab. The mean and maximum dose were calculated for the Planning Target Volume (PTV1),more » parotids, brainstem, and spinal cord and then compared to the original baseline plan. The D1cc was also calculated for the spinal cord and brainstem. Patient average results were compared across institutions. Results: Introduced gantry angle errors had the smallest effect of dose, no tolerances were exceeded for one institution, and the second institutions VMAT plans were only exceeded for gantry angle of ±5° affecting different sided parotids by 14–18%. PTV1, brainstem and spinal cord tolerances were exceeded for collimator angles of ±5 degrees, MLC shifts and MLC field sizes of ±1 and beyond, at the first institution. At the second institution, sensitivity to errors was marginally higher for some errors including the collimator error producing doses exceeding tolerances above ±2 degrees, and marginally lower with tolerances exceeded above MLC shifts of ±2. The largest differences occur with MLC field sizes, with both institutions reporting exceeded tolerances, for all introduced errors (±1 and beyond). Conclusion: The plan robustness for VMAT nasopharynx plans has been demonstrated. Gantry errors have the least impact on patient doses, however MLC field sizes exceed tolerances even with relatively low introduced errors and also produce the largest errors. This was consistent across both departments. The authors acknowledge funding support from the NSW Cancer Council.« less
Brushed permanent magnet DC MLC motor operation in an external magnetic field.
Yun, J; St Aubin, J; Rathee, S; Fallone, B G
2010-05-01
Linac-MR systems for real-time image-guided radiotherapy will utilize the multileaf collimators (MLCs) to perform conformal radiotherapy and tumor tracking. The MLCs would be exposed to the external fringe magnetic fields of the linac-MR hybrid systems. Therefore, an experimental investigation of the effect of an external magnetic field on the brushed permanent magnet DC motors used in some MLC systems was performed. The changes in motor speed and current were measured for varying external magnetic field strengths up to 2000 G generated by an EEV electromagnet. These changes in motor characteristics were measured for three orientations of the motor in the external magnetic field, mimicking changes in motor orientations due to installation and/or collimator rotations. In addition, the functionality of the associated magnetic motor encoder was tested. The tested motors are used with the Varian 120 leaf Millennium MLC (Maxon Motor half leaf and full leaf motors) and the Varian 52 leaf MKII MLC (MicroMo Electronics leaf motor) including a carriage motor (MicroMo Electronics). In most cases, the magnetic encoder of the motors failed prior to any damage to the gearbox or the permanent magnet motor itself. This sets an upper limit of the external magnetic field strength on the motor function. The measured limits of the external magnetic fields were found to vary by the motor type. The leaf motor used with a Varian 52 leaf MKII MLC system tolerated up to 450 +/- 10 G. The carriage motor tolerated up to 2000 +/- 10 G field. The motors used with the Varian 120 leaf Millennium MLC system were found to tolerate a maximum of 600 +/- 10 G. The current Varian MLC system motors can be used for real-time image-guided radiotherapy coupled to a linac-MR system, provided the fringe magnetic fields at their locations are below the determined tolerance levels. With the fringe magnetic fields of linac-MR systems expected to be larger than the tolerance levels determined, some form of magnetic shielding would be required.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Lebron, S; Yan, G; Li, J
2016-06-15
Purpose: To develop an accurate and quick multileaf collimator (MLC) calibration and quality assurance technique using an electronic portal imaging device (EPID) Methods: The MLC models used include the MLCi and Agility (Elekta Ltd). This technique consists of two 22(L)x10(W) cm{sup 2} fields with 0{sup 0} and 180{sup 0} collimator angles centered to an offset EPID. The MLC opening is estimated by calculating the profile at the image’s center in the image’s horizontal direction. Scans in the image’s vertical direction were calculated every 20 pixels in the inner 70% of estimated MLC opening. The profiles’ edges were fitted with linearmore » equations to determine the image’s rotation angle. Then, crossline profiles were scanned at the center of each leaf taking into account the leaf’s width at isocenter and the rotation angle. The profiles’ edges determine the location of the leaves’ edges and these were subtracted from the reference leaf’s position in order to determine the relative leaf offsets. The edge location of all profiles was determined by using the parameterized gradient of the penumbra region. The technique was tested against an established diode array-based method, and for different MLC systems, patterns, gantry angles, days, energies, beam modalities and MLC openings. Results: The differences between the proposed and established methods were 0.26±0.19mm. The leaf offsets’ deviation was <0.3mm (5 months period). For pattern fields, the differences between predetermined and calculated offsets were 0.18±0.18mm. The leaf offset deviation of measurements with different energies and MLC openings were <0.1mm and <0.3mm, respectively. The differences between offsets of FF and FFF beams were 0.01±0.02mm (<0.07mm). The differences between the offsets at different gantry angles were 0.08±0.15mm. Conclusion: The proposed method proved to be accurate and efficient in calculating the relative leaf offsets. Parameterized field edge is essential to obtain accurate result by eliminating the noise from EPID.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Eley, John G.; Hogstrom, Kenneth R.; Matthews, Kenneth L.
2011-12-15
Purpose: The purpose of this work was to investigate the potential of discrete Gaussian edge feathering of the higher energy electron fields for improving abutment dosimetry in the planning volume when using an electron multileaf collimator (eMLC) to deliver segmented-field electron conformal therapy (ECT). Methods: A discrete (five-step) Gaussian edge spread function was used to match dose penumbras of differing beam energies (6-20 MeV) at a specified depth in a water phantom. Software was developed to define the leaf eMLC positions of an eMLC that most closely fit each electron field shape. The effect of 1D edge feathering of themore » higher energy field on dose homogeneity was computed and measured for segmented-field ECT treatment plans for three 2D PTVs in a water phantom, i.e., depth from the water surface to the distal PTV surface varied as a function of the x-axis (parallel to leaf motion) and remained constant along the y-axis (perpendicular to leaf motion). Additionally, the effect of 2D edge feathering was computed and measured for one radially symmetric, 3D PTV in a water phantom, i.e., depth from the water surface to the distal PTV surface varied as a function of both axes. For the 3D PTV, the feathering scheme was evaluated for 0.1-1.0-cm leaf widths. Dose calculations were performed using the pencil beam dose algorithm in the Pinnacle{sup 3} treatment planning system. Dose verification measurements were made using a prototype eMLC (1-cm leaf width). Results: 1D discrete Gaussian edge feathering reduced the standard deviation of dose in the 2D PTVs by 34, 34, and 39%. In the 3D PTV, the broad leaf width (1 cm) of the eMLC hindered the 2D application of the feathering solution to the 3D PTV, and the standard deviation of dose increased by 10%. However, 2D discrete Gaussian edge feathering with simulated eMLC leaf widths of 0.1-0.5 cm reduced the standard deviation of dose in the 3D PTV by 33-28%, respectively. Conclusions: A five-step discrete Gaussian edge spread function applied in 2D improves the abutment dosimetry but requires an eMLC leaf resolution better than 1 cm.« less
Role of Rac1 in Escherichia coli K1 invasion of human brain microvascular endothelial cells.
Rudrabhatla, Rajyalakshmi S; Selvaraj, Suresh K; Prasadarao, Nemani V
2006-02-01
Escherichia coli K1 invasion of human brain microvascular endothelial cells (HBMEC) requires the reorganization of host cytoskeleton at the sites of bacterial entry. Both actin and myosin constitute the cytoskeletal architecture. We have previously shown that myosin light chain (MLC) phosphorylation by MLC kinase is regulated during E. coli invasion by an upstream kinase, p21-activated kinase 1 (PAK1), which is an effector protein of Rac and Cdc42 GTPases, but not of RhoA. Here, we report that the binding of only Rac1 to PAK1 decreases in HBMEC upon infection with E. coli K1, which resulted in increased phosphorylation of MLC. Overexpression of a constitutively active (cAc) form of Rac1 in HBMEC blocked the E. coli invasion significantly, whereas overexpression of a dominant negative form had no effect. Increased PAK1 phosphorylation was observed in HBMEC expressing cAc-Rac1 with a concomitant reduction in the phosphorylation of MLC. Immunocytochemistry studies demonstrated that the inhibition of E. coli invasion into cAc-Rac1/HBMEC is due to lack of phospho-MLC recruitment to the sites of E. coli entry. Taken together the data suggest that E. coli modulates the binding of Rac1, but not Cdc42, to PAK1 during the invasion of HBMEC.
Steel, Jared; Stewart, Allan; Satory, Philip
2009-09-01
Matching the penumbra of a 6 MeV electron beam to the penumbra of a 6 MV photon beam is a dose optimization challenge, especially when the electron beam is applied from an extended source-to-surface distance (SSD), as in the case of some head and neck treatments. Traditionally low melting point alloy blocks have been used to define the photon beam shielding over the spinal cord region. However, these are inherently time consuming to construct and employ in the clinical situation. Multileaf collimators (MLCs) provide a fast and reproducible shielding option but generate geometrically nonconformal approximations to the desired beam edge definition. The effects of substituting Cerrobend for the MLC shielding mode in the context of beam matching with extended-SSD electron beams are the subject of this investigation. Relative dose beam data from a Varian EX 2100 linear accelerator were acquired in a water tank under the 6 MeV electron beam at both standard and extended-SSD and under the 6 MV photon beam defined by Cerrobend and a number of MLC stepping regimes. The effect of increasing the electron beam SSD on the beam penumbra was assessed. MLC stepping was also assessed in terms of the effects on both the mean photon beam penumbra and the intraleaf dose-profile nonuniformity relative to the MLC midleaf. Computational techniques were used to combine the beam data so as to simulate composite relative dosimetry in the water tank, allowing fine control of beam abutment gap variation. Idealized volumetric dosimetry was generated based on the percentage depth-dose data for the beam modes and the abutment geometries involved. Comparison was made between each composite dosimetry dataset and the relevant ideal dosimetry dataset by way of subtraction. Weighted dose-difference volume histograms (DDVHs) were produced, and these, in turn, summed to provide an overall dosimetry score for each abutment and shielding type/angle combination. Increasing the electron beam SSD increased the penumbra width (defined as the lateral distance of the 80% and 20% isodose contours) by 8-10 mm at the depths of 10-20 mm. Mean photon beam penumbra width increased with increased MLC stepping, and the mean MLC penumbra was approximately 1.5 times greater than that across the corresponding Cerrobend shielding. Intraleaf dose discrepancy in the direction orthogonal to the beam edge also increased with MLC stepping. The weighted DDVH comparison techniques allowed the composite dosimetry resulting from the interplay of the abovementioned variables to be ranked. The MLC dosimetry ranked as good or better than that resulting from beam matching with Cerrobend for all except large field overlaps (-2.5 mm gap). The results for the linear-weighted DDVH comparison suggest that optimal MLC abutment dosimetry results from an optical surface gap of around 1 +/- 0.5 mm. Furthermore, this appears reasonably lenient to abutment gap variation, such as that arising from uncertainty in beam markup or other setup errors.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Riis, Hans L.; Zimmermann, Sune J.; Hjelm-Hansen, Mogens
Purpose: The delivery of high quality stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT) treatments to the patient requires knowledge of the position of the isocenter to submillimeter accuracy. To meet the requirements the deviation between the radiation and mechanical isocenters must be less than 1 mm. The use of add-on micromultileaf collimators ({mu}MLCs) in SRS and SRT is an additional challenge to the anticipated high-level geometric and dosimetric accuracy of the treatment. The aim of this work was to quantify the gantry excursions during rotation with and without an add-on {mu}MLC attached to the gantry head. In addition, the shiftmore » in the position of the isocenter and its correlation to the kV beam center of the cone-beam CT system was included in the study. Methods: The quantification of the gantry rotational performance was done using a pointer supported by an in-house made rigid holder attached to the gantry head of the accelerator. The pointer positions were measured using a digital theodolite. To quantify the effect of an {mu}MLC of 50 kg, the measurements were repeated with the {mu}MLC attached to the gantry head. The displacement of the isocenter due to an add-on {mu}MLC of 50 kg was also investigated. In case of the pointer measurement the {mu}MLC was simulated by weights attached to the gantry head. A method of least squares was applied to determine the position and displacement of the mechanical isocenter. Additionally, the displacement of the radiation isocenter was measured using a ball-bearing phantom and the electronic portal image device system. These measurements were based on 8 MV photon beams irradiated onto the ball from the four cardinal angles and two opposed collimator angles. The measurements and analysis of the data were carried out automatically using software delivered by the manufacturer. Results: The displacement of the mechanical isocenter caused by a 50 kg heavy {mu}MLC was found to be (-0.01 {+-} 0.05, -0.10 {+-} 0.03, -0.26 {+-} 0.05) mm in lateral, longitudinal, and vertical direction, respectively. Similarly, the displacement of the radiation isocenter was found to be (0.00 {+-} 0.03, -0.08 {+-} 0.06, -0.32 {+-} 0.02) mm. Good agreement was found between the displacement of the two isocenters. A displacement of the kV cone-beam CT beam center due to the attached weight of 50 kg could not be detected. Conclusions: General characteristics of the gantry arm excursions and displacements caused by an add-on {mu}MLC have been reported. A 50 kg heavy add-on {mu}MLC results in a isocenter displacement downward of 0.26-0.32 mm. The authors recommend that the beam center of the kV cone-beam CT image system should be matched to the isocenter related to the weight of the {mu}MLC. Consequently, the imperfections in isocenter localizations are transferred to the conventional radiotherapy where the clinical consequences of uncertainties in the submillimeter regime are negligible.« less
Well 14-2 Logs and Data: Roosevelt Hot Spring Area, Utah (Utah FORGE)
Joe Moore
2016-03-03
This is a compilation of logs and data from Well 14-2 in the Roosevelt Hot Springs area in Utah. This well is also in the Utah FORGE study area. The file is in a compressed .zip format and there is a data inventory table (Excel spreadsheet) in the root folder that is a guide to the data that is accessible in subfolders.
Well 52-21 Logs and Data: Roosevelt Hot Spring Area, Utah (Utah FORGE)
Joe Moore
2016-03-03
This is a compilation of logs and data from Well 52-21 in the Roosevelt Hot Springs area in Utah. This well is also in the Utah FORGE study area. The file is in a compressed .zip format and there is a data inventory table (Excel spreadsheet) in the root folder that is a guide to the data that is accessible in subfolders.
Well 82-33 Logs and Data: Roosevelt Hot Spring Area, Utah (Utah FORGE)
Joe Moore
2016-03-03
This is a compilation of logs and data from Well 82-33 in the Roosevelt Hot Springs area in Utah. This well is also in the Utah FORGE study area. The file is in a compressed .zip format and there is a data inventory table (Excel spreadsheet) in the root folder that is a guide to the data that is accessible in subfolders.
Well Acord 1-26 Logs and Data: Roosevelt Hot Spring Area, Utah (Utah FORGE)
DOE Office of Scientific and Technical Information (OSTI.GOV)
Joe Moore
This is a compilation of logs and data from Well Acord 1-26 in the Roosevelt Hot Springs area in Utah. This well is also in the Utah FORGE study area. The file is in a compressed .zip format and there is a data inventory table (Excel spreadsheet) in the root folder that is a guide to the data that is accessible in subfolders.
Patterns of usage for a Web-based clinical information system.
Chen, Elizabeth S; Cimino, James J
2004-01-01
Understanding how clinicians are using clinical information systems to assist with their everyday tasks is valuable to the system design and development process. Developers of such systems are interested in monitoring usage in order to make enhancements. System log files are rich resources for gaining knowledge about how the system is being used. We have analyzed the log files of our Web-based clinical information system (WebCIS) to obtain various usage statistics including which WebCIS features are frequently being used. We have also identified usage patterns, which convey how the user is traversing the system. We present our method and these results as well as describe how the results can be used to customize menus, shortcut lists, and patient reports in WebCIS and similar systems.
Simulation Control Graphical User Interface Logging Report
NASA Technical Reports Server (NTRS)
Hewling, Karl B., Jr.
2012-01-01
One of the many tasks of my project was to revise the code of the Simulation Control Graphical User Interface (SIM GUI) to enable logging functionality to a file. I was also tasked with developing a script that directed the startup and initialization flow of the various LCS software components. This makes sure that a software component will not spin up until all the appropriate dependencies have been configured properly. Also I was able to assist hardware modelers in verifying the configuration of models after they have been upgraded to a new software version. I developed some code that analyzes the MDL files to determine if any error were generated due to the upgrade process. Another one of the projects assigned to me was supporting the End-to-End Hardware/Software Daily Tag-up meeting.
Risser, Dennis W.; Williams, John H.; Hand, Kristen L.; Behr, Rose-Anna; Markowski, Antonette K.
2013-01-01
Open-File Miscellaneous Investigation 13–01.1 presents the results of geohydrologic investigations on a 1,664-foot-deep core hole drilled in the Bradford County part of the Gleason 7.5-minute quadrangle in north-central Pennsylvania. In the text, the authors discuss their methods of investigation, summarize physical and analytical results, and place those results in context. Four appendices include (1) a full description of the core in an Excel worksheet; (2) water-quality and core-isotope analytical results in Excel workbooks; (3) geophysical logs in LAS and PDF files, and an Excel workbook containing attitudes of bedding and fractures calculated from televiewer logs; and (4) MP4 clips from the downhole video at selected horizons.
Minimal-Level Cataloging: A Look at the Issues--A Symposium.
ERIC Educational Resources Information Center
Horny, Karen L.; And Others
1986-01-01
Six articles on pros and cons of minimal-level cataloging (MLC) highlight patron access and browsability, time and cost savings, network and bibliographic utility cooperation, standards, alternatives to MLC, levels of records, online catalogs, experimental cataloging program at University of Michigan, and experiences at Kent State University…
Zebra: A striped network file system
NASA Technical Reports Server (NTRS)
Hartman, John H.; Ousterhout, John K.
1992-01-01
The design of Zebra, a striped network file system, is presented. Zebra applies ideas from log-structured file system (LFS) and RAID research to network file systems, resulting in a network file system that has scalable performance, uses its servers efficiently even when its applications are using small files, and provides high availability. Zebra stripes file data across multiple servers, so that the file transfer rate is not limited by the performance of a single server. High availability is achieved by maintaining parity information for the file system. If a server fails its contents can be reconstructed using the contents of the remaining servers and the parity information. Zebra differs from existing striped file systems in the way it stripes file data: Zebra does not stripe on a per-file basis; instead it stripes the stream of bytes written by each client. Clients write to the servers in units called stripe fragments, which are analogous to segments in an LFS. Stripe fragments contain file blocks that were written recently, without regard to which file they belong. This method of striping has numerous advantages over per-file striping, including increased server efficiency, efficient parity computation, and elimination of parity update.
ERIC Educational Resources Information Center
Bennett-Abney, Cheryl
2001-01-01
Three organizational tools for counselors are described: three-ring binder for notes, forms, and schedules; daily log of time and activities; and a tickler file with tasks arranged by days of the week. (SK)
DOE Office of Scientific and Technical Information (OSTI.GOV)
Shiinoki, T; Hanazawa, H; Park, S
2015-06-15
Purpose: We aim to achieve new four-dimensional radiotherapy (4DRT) using the next generation real-time tumor-tracking (RTRT) system and flattening-filter-free techniques. To achieve new 4DRT, it is necessary to understand the respiratory motion of tumor. The purposes of this study were: 1.To develop the respiratory motion analysis tool using log files. 2.To evaluate the reproducibility of tumor motion probability distribution function (PDF) during stereotactic body RT (SBRT) of lung tumor. Methods: Seven patients having fiducial markers closely implanted to the lung tumor were enrolled in this study. The positions of fiducial markers were measured using the RTRT system (Mitsubishi Electronics Co.,more » JP) and recorded as two types of log files during the course of SBRT. For each patients, tumor motion range and tumor motion PDFs in left-right (LR), anterior-posterior (AP) and superior-inferior (SI) directions were calculated using log files of all beams per fraction (PDFn). Fractional PDF reproducibility (Rn) was calculated as Kullback-Leibler (KL) divergence between PDF1 and PDFn of tumor motion. The mean of Rn (Rm) was calculated for each patient and correlated to the patient’s mean tumor motion range (Am). The change of Rm during the course of SBRT was also evluated. These analyses were performed using in-house developed software. Results: The Rm were 0.19 (0.07–0.30), 0.14 (0.07–0.32) and 0.16 (0.09–0.28) in LR, AP and SI directions, respectively. The Am were 5.11 mm (2.58–9.99 mm), 7.81 mm (2.87–15.57 mm) and 11.26 mm (3.80–21.27 mm) in LR, AP and SI directions, respectively. The PDF reproducibility decreased as the tumor motion range increased in AP and SI direction. That decreased slightly through the course of RT in SI direction. Conclusion: We developed the respiratory motion analysis tool for 4DRT using log files and quantified the range and reproducibility of respiratory motion for lung tumors.« less
2013-03-01
the /bin, /sbin, /etc, /var/log, /home, /proc, /root, /dev, /tmp, and /lib directories • Describe the purpose of the /etc/shadow and /etc/ passwd ...UNLIMITED 19 2.6.2 /etc/ passwd and /etc/shadow The /etc/shadow file didn’t exist on early Linux distributions. Originally only root could access the...etc/ passwd file, which stored user names, user configuration information, and passwords. However, when common programs such as ls running under
DOE Office of Scientific and Technical Information (OSTI.GOV)
Heath, Emily; Seuntjens, Jan; Sheikh-Bagheri, Daryoush
2004-10-01
In this work we dosimetrically evaluated the clinical implementation of a commercial Monte Carlo treatment planning software (PEREGRINE, North American Scientific, Cranberry Township, PA) intended for quality assurance (QA) of intensity modulated radiation therapy treatment plans. Dose profiles calculated in homogeneous and heterogeneous phantoms using this system were compared to both measurements and simulations using the EGSnrc Monte Carlo code for the 6 MV beam of a Varian CL21EX linear accelerator. For simple jaw-defined fields, calculations agree within 2% of the dose at d{sub max} with measurements in homogeneous phantoms with the exception of the buildup region where the calculationsmore » overestimate the dose by up to 8%. In heterogeneous lung and bone phantoms the agreement is within 3%, on average, up to 5% for a 1x1 cm{sup 2} field. We tested two consecutive implementations of the MLC model. After matching the calculated and measured MLC leakage, simulations of static and dynamic MLC-defined fields using the most recent MLC model agreed to within 2% with measurements.« less
NASA Astrophysics Data System (ADS)
Tyagi, N.; Curran, B. H.; Roberson, P. L.; Moran, J. M.; Acosta, E.; Fraass, B. A.
2008-02-01
IMRT often requires delivering small fields which may suffer from electronic disequilibrium effects. The presence of heterogeneities, particularly low-density tissues in patients, complicates such situations. In this study, we report on verification of the DPM MC code for IMRT treatment planning in heterogeneous media, using a previously developed model of the Varian 120-leaf MLC. The purpose of this study is twofold: (a) design a comprehensive list of experiments in heterogeneous media for verification of any dose calculation algorithm and (b) verify our MLC model in these heterogeneous type geometries that mimic an actual patient geometry for IMRT treatment. The measurements have been done using an IMRT head and neck phantom (CIRS phantom) and slab phantom geometries. Verification of the MLC model has been carried out using point doses measured with an A14 slim line (SL) ion chamber inside a tissue-equivalent and a bone-equivalent material using the CIRS phantom. Planar doses using lung and bone equivalent slabs have been measured and compared using EDR films (Kodak, Rochester, NY).
NASA Astrophysics Data System (ADS)
Kim, Seung-Tae; Cho, Won-Ju
2018-01-01
We fabricated a resistive random access memory (ReRAM) device on a Ti/AlO x /Pt structure with solution-processed AlO x switching layer using microwave irradiation (MWI), and demonstrated multi-level cell (MLC) operation. To investigate the effect of MWI power on the MLC characteristics, post-deposition annealing was performed at 600-3000 W after AlO x switching layer deposition, and the MLC operation was compared with as-deposited (as-dep) and conventional thermally annealing (CTA) treated devices. All solution-processed AlO x -based ReRAM devices exhibited bipolar resistive switching (BRS) behavior. We found that these devices have four-resistance states (2 bits) of MLC operation according to the modulation of the high-resistance state (HRSs) through reset voltage control. Particularly, compared to the as-dep and CTA ReRAM devices, the MWI-treated ReRAM devices showed a significant increase in the memory window and stable endurance for multi-level operation. Moreover, as the MWI power increased, excellent MLC characteristics were exhibited because the resistance ratio between each resistance state was increased. In addition, it exhibited reliable retention characteristics without deterioration at 25 °C and 85 °C for 10 000 s. Finally, the relationship between the chemical characteristics of the solution-processed AlO x switching layer and BRS-based multi-level operation according to the annealing method and MWI power was investigated using x-ray photoelectron spectroscopy.
Leeuw, T; Pette, D
1996-01-01
Skeletal muscle fibers are versatile entities, capable of changing their phenotype in response to altered functional demands. In the present study, fast-to-slow fiber type transitions were induced in rabbit tibialis anterior (fA) muscles by chronic low-frequency stimulation (CLFS). The time course of changes in relative protein concentrations of fast and slow myosin light chain (MLC) isoforms and changes in their relative synthesis rates by in vivo labeling with [35S]methionine were followed during stimulation periods of up to 60 days. Generally, relative synthesis rates and protein concentrations changed in parallel; i.e., fast isoforms decreased and slow isoforms increased. MLC3f, however, which turns over at a higher rate than the other light chains, exhibited a conspicuous discrepancy between a markedly reduced relative synthesis but only a moderate decrease in protein amount during the initial 2 weeks of CLFS. Apparently, MLC3f is regulated independent of MLC1f, with protein degradation playing an important role in its regulation. The exchange of fast MLC isoforms with their slow counterparts seemed to correspond to the ultimate fast-to-slow (MHCIIa-->MHCI) transition at the MHC level. However, due to an earlier onset of the fast-to-slow transition of the regulatory light chain and the delayed fast-to-slow exchange of the alkali light chains, a spectrum of hybrid isomyosins composed of fast and slow light and heavy chains must have existed transiently in transforming fibers. Such hybrid isomyosins appeared to be restricted to MHCIIa- and MHCI-based combinations. In conclusion, fiber type specific programs that normally coordinate the expression of myofibrillar protein isoforms seem to be maintained during fiber type transitions. Possible differences in post-transcriptional regulation may result in the transient accumulation of atypical combinations of fast and slow MLC and MHC isoforms, giving rise to the appearance of hybrid fibers under the conditions of forced fiber type conversion.
Bae, Yun Jung; Yoon, Young Il; Yoon, Tae-Jong; Lee, Hak Jong
2016-01-01
To evaluate the effectiveness of ultrasound and microbubble-liposome complex (MLC)-mediated delivery of siRNA and doxorubicin into prostate cancer cells and its therapeutic capabilities both in vitro and in vivo. Microbubble-liposome complexes conjugated with anti-human epidermal growth factor receptor type 2 (Her2) antibodies were developed to target human prostate cancer cell lines PC-3 and LNCaP. Intracellular delivery of MLC was observed by confocal microscopy. We loaded MLC with survivin-targeted small interfering RNA (siRNA) and doxorubicin, and delivered it into prostate cancer cells. The release of these agents was facilitated by ultrasound application. Cell viability was analyzed by MTT assay after the delivery of siRNA and doxorubicin. Survivin-targeted siRNA loaded MLC was delivered into the xenograft mouse tumor model. Western blotting was performed to quantify the expression of survivin in vivo. Confocal microscopy demonstrated substantial intracellular uptake of MLCs in LNCaP, which expresses higher levels of Her2 than PC-3. The viability of LNCaP cells was significantly reduced after the delivery of MLCs loaded with siRNA and doxorubicin (85.0 ± 2.9%), which was further potentiated by application of ultrasound (55.0 ± 3.5%, p = 0.009). Survivin expression was suppressed in vivo in LNCaP tumor xenograft model following the ultrasound and MLC-guided delivery of siRNA (77.4 ± 4.90% to 36.7 ± 1.34%, p = 0.027). Microbubble-liposome complex can effectively target prostate cancer cells, enabling intracellular delivery of the treatment agents with the use of ultrasound. Ultrasound and MLC-mediated delivery of survivin-targeted siRNA and doxorubicin can induce prostate cell apoptosis and block survivin expression in vitro and in vivo.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Tien, C; Brewer, M; Studenski, M
Purpose: Dynamic-jaw tracking maximizes the area blocked by both jaw and MLC in RapidArc. We developed a method to quantify jaw tracking. Methods: An Eclipse Scripting API (ESAPI) was used to export beam parameters for each arc’s control points. The specific beam parameters extracted were: gantry angle, control point number, meterset, x-jaw positions, y-jaw positions, MLC bank-number, MLC leaf-number, and MLC leaf-position. Each arc contained 178 control points with 120 MLC positions. MATLAB routines were written to process these parameters in order to calculate both the beam aperture (unblocked) size for each control point. An average aperture size was weightedmore » by meterset. Jaw factor was defined as the ratio between dynamic-jaw to static-jaw aperture size. Jaw factor was determined for forty retrospectively replanned patients treated with static-jaw delivery sites including lung, brain, prostate, H&N, rectum, and bladder. Results: Most patients had multiple arcs and reduced-field boosts, resulting in 151 fields. Of these, the lowest (0.4722) and highest (0.9622) jaw factor was observed in prostate and rectal cases, respectively. The median jaw factor was 0.7917 meaning there is the potential unincreased blocking by 20%. Clinically, the dynamic-jaw tracking represents an area surrounding the target which would receive MLC-only leakage transmission of 1.68% versus 0.1% with jaws. Jaw-tracking was more pronounced at areas farther from the target. In prostate patients, the rectum and bladder had 5.5% and 6.3% lower mean dose, respectively; the structures closer to the prostate such as the rectum and bladder both had 1.4% lower mean dose. Conclusion: A custom ESAPI script was coupled with a MATLAB routine in order to extract beam parameters from static-jaw plans and their replanned dynamic-jaw deliveries. The effects were quantified using jaw factor which is the ratio between the meterset weighted aperture size for dynamic-jaw fields versus static-jaw fields.« less
SU-E-T-179: Clinical Impact of IMRT Failure Modes at Or Near TG-142 Tolerance Criteria Levels
DOE Office of Scientific and Technical Information (OSTI.GOV)
Faught, J Tonigan; Balter, P; Johnson, J
2015-06-15
Purpose: Quantitatively assess the clinical impact of 11 critical IMRT dose delivery failure modes. Methods: Eleven step-and-shoot IMRT failure modes (FMs) were introduced into twelve Pinnacle v9.8 treatment plans. One standard and one highly modulated plan on the IROC IMRT phantom and ten previous H&N patient treatment plans were used. FMs included physics components covered by basic QA near tolerance criteria levels (TG-142) such as beam energy, MLC positioning, and MLC modeling. Resultant DVHs were compared to those of failure-free plans and the severity of plan degradation was assessed considering PTV coverage and OAR and normal tissue tolerances and usedmore » for FMEA severity scoring. Six of these FMs were physically simulated and phantom irradiations performed. TLD and radiochromic film results are used for comparison to treatment planning studies. Results: Based on treatment planning studies, the largest clinical impact from the phantom cases was induced by 2 mm systematic MLC shift in one bank with the combination of a D95% target under dose near 16% and OAR overdose near 8%. Cord overdoses of 5%–11% occurred with gantry angle, collimator angle, couch angle, MLC leaf end modeling, and MLC transmission and leakage modeling FMs. PTV coverage and/or OAR sparing was compromised in all FMs introduced in phantom plans with the exception of CT number to electron density tables, MU linearity, and MLC tongue-and-groove modeling. Physical measurements did not entirely agree with treatment planning results. For example, symmetry errors resulted in the largest physically measured discrepancies of up to 3% in the PTVs while a maximum of 0.5% deviation was seen in the treatment planning studies. Patient treatment plan study results are under analysis. Conclusion: Even in the simplistic anatomy of the IROC phantom, some basic physics FMs, just outside of TG-142 tolerance criteria, appear to have the potential for large clinical implications.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Li, JS; Fan, J; Ma, C-M
Purpose: To improve the treatment efficiency and capabilities for full-body treatment, a robotic radiosurgery system has equipped with a multileaf collimator (MLC) to extend its accuracy and precision to radiation therapy. To model the MLC and include it in the Monte Carlo patient dose calculation is the goal of this work. Methods: The radiation source and the MLC were carefully modeled to consider the effects of the source size, collimator scattering, leaf transmission and leaf end shape. A source model was built based on the output factors, percentage depth dose curves and lateral dose profiles measured in a water phantom.more » MLC leaf shape, leaf end design and leaf tilt for minimizing the interleaf leakage and their effects on beam fluence and energy spectrum were all considered in the calculation. Transmission/leakage was added to the fluence based on the transmission factors of the leaf and the leaf end. The transmitted photon energy was tuned to consider the beam hardening effects. The calculated results with the Monte Carlo implementation was compared with measurements in homogeneous water phantom and inhomogeneous phantoms with slab lung or bone material for 4 square fields and 9 irregularly shaped fields. Results: The calculated output factors are compared with the measured ones and the difference is within 1% for different field sizes. The calculated dose distributions in the phantoms show good agreement with measurements using diode detector and films. The dose difference is within 2% inside the field and the distance to agreement is within 2mm in the penumbra region. The gamma passing rate is more than 95% with 2%/2mm criteria for all the test cases. Conclusion: Implementation of Monte Carlo dose calculation for a MLC equipped robotic radiosurgery system is completed successfully. The accuracy of Monte Carlo dose calculation with MLC is clinically acceptable. This work was supported by Accuray Inc.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Kong, X; Harris, J; Spitznagel, D
2015-06-15
Purpose: To analyze the radiation transmission of the Agility MLC and make recommendation based on the MLC dosimetric characteristics for SRT, SBRT and VMAT planning Method and Materials: Agility MLC is the newest model from Elekta and has no back up diaphragm behind leaves for this generation. Leaves are single focused with rounded end; composed of leafs each 0.5cm wide, 9cm thick, constructed from tungsten alloy and provide low transmission <0.5%. Total radiation transmission from leaves and diaphragm is <0.13%. A 360degree arc was generated using iCom tools; leaves were programmed closed while keeping the diaphragm fully open to maximizemore » the MLC transmission effect. Gafchromic EBT films were sandwiched between 4cm of solid water and situated at midplane to take dose measurement. 5000MU was delivered using 6MV VersaHD, various collimator angles, and a 5cm central axis offset was tested also. Films were scanned with Epson 10000XL scanner and analyzed using DoseLab Pro. Results: Due to the rounded leaf end and nature of rotation therapy, dose accumulation through the leaf gap is significant. By offsetting the leaf gap from central axis, this accumulation can be greatly reduced. There are dark bands showing accumulation of interleaf transmission which is improved by increasing collimator angle from 0 to 45dgree. However for 45 degree, in most cases, there are larger volumes sweeping under MLC alone, which needs considered planning. Conclusions: While inter-leaf leakage is minimized by using collimator angles greater than 0 degrees, the location of the leaf gap must also be managed. The leaf gap position becomes critically important when the treatment area is off axis such is the case when more than one PTV is being treated. With VMAT for SRT, SBRT becoming a more popular planning technique, special attention needs to be paid when initially setting up the field geometry.« less
SU-E-T-610: Comparison of Treatment Times Between the MLCi and Agility Multileaf Collimators
DOE Office of Scientific and Technical Information (OSTI.GOV)
Ramsey, C; Bowling, J
2014-06-01
Purpose: The Agility is a new 160-leaf MLC developed by Elekta for use in their Infinity and Versa HD linacs. As compared to the MLCi, the Agility increased the maximum leaf speed from 2 cm/s to 3.5 cm/s, and the maximum primary collimator speed from 1.5 cm/s to 9.0 cm/s. The purpose of this study was to determine if the Agility MLC resulted in improved plan quality and/or shorter treatment times. Methods: An Elekta Infinity that was originally equipped with a 80 leaf MLCi was upgraded to an 160 leaf Agility. Treatment plan quality was evaluated using the Pinnacle planningmore » system with SmartArc. Optimization was performed once for the MLCi and once for the Agility beam models using the same optimization parameters and the same number of iterations. Patient treatment times were measured for all IMRT, VMAT, and SBRT patients treated on the Infinity with the MLCi and Agility MLCs. Treatment times were extracted from the EMR and measured from when the patient first walked into the treatment room until exiting the treatment room. Results: 11,380 delivery times were measured for patients treated with the MLCi, and 1,827 measurements have been made for the Agility MLC. The average treatment times were 19.1 minutes for the MLCi and 20.8 minutes for the Agility. Using a t-test analysis, there was no difference between the two groups (t = 0.22). The dose differences between patients planned with the MLCi and the Agility MLC were minimal. For example, the dose difference for the PTV, GTV, and cord for a head and neck patient planned using Pinnacle were effectively equivalent. However, the dose to the parotid glands was slightly worse with the Agility MLC. Conclusion: There was no statistical difference in treatment time, or any significant dosimetric difference between the Agility MLC and the MLCi.« less
Kielar, Kayla N; Mok, Ed; Hsu, Annie; Wang, Lei; Luxton, Gary
2012-10-01
The dosimetric leaf gap (DLG) in the Varian Eclipse treatment planning system is determined during commissioning and is used to model the effect of the rounded leaf-end of the multileaf collimator (MLC). This parameter attempts to model the physical difference between the radiation and light field and account for inherent leakage between leaf tips. With the increased use of single fraction high dose treatments requiring larger monitor units comes an enhanced concern in the accuracy of leakage calculations, as it accounts for much of the patient dose. This study serves to verify the dosimetric accuracy of the algorithm used to model the rounded leaf effect for the TrueBeam STx, and describes a methodology for determining best-practice parameter values, given the novel capabilities of the linear accelerator such as flattening filter free (FFF) treatments and a high definition MLC (HDMLC). During commissioning, the nominal MLC position was verified and the DLG parameter was determined using MLC-defined field sizes and moving gap tests, as is common in clinical testing. Treatment plans were created, and the DLG was optimized to achieve less than 1% difference between measured and calculated dose. The DLG value found was tested on treatment plans for all energies (6 MV, 10 MV, 15 MV, 6 MV FFF, 10 MV FFF) and modalities (3D conventional, IMRT, conformal arc, VMAT) available on the TrueBeam STx. The DLG parameter found during the initial MLC testing did not match the leaf gap modeling parameter that provided the most accurate dose delivery in clinical treatment plans. Using the physical leaf gap size as the DLG for the HDMLC can lead to 5% differences in measured and calculated doses. Separate optimization of the DLG parameter using end-to-end tests must be performed to ensure dosimetric accuracy in the modeling of the rounded leaf ends for the Eclipse treatment planning system. The difference in leaf gap modeling versus physical leaf gap dimensions is more pronounced in the more recent versions of Eclipse for both the HDMLC and the Millennium MLC. Once properly commissioned and tested using a methodology based on treatment plan verification, Eclipse is able to accurately model radiation dose delivered for SBRT treatments using the TrueBeam STx.
DOE Office of Scientific and Technical Information (OSTI.GOV)
You, T; Dang, J; Dai, C
2015-06-15
Purpose: To evaluate dosimetric impact of spinal SBRT when MLC leaf positions deviate from planning positions for different energies and doserates. Methods and Materials: 18 localized spinal metastases patients were selected for SBRT using IMRT planning with 9 posterior beams delivered at gantry angles ranging between 100°–260°. A modern linear accelerator(Varian Turebeam STX with HDMLC 2.5 mm thick leaf at isocenter) IMRT plans were generated using both 6X and 6X-FFF(Flattening filter free) beams with a nominal prescription dose of 6 Gy/fraction to PTV. Doserates ranging from 200–600 MU/min for 6X and 400–1400 MU/min for 6X-FFF, with 200 increments were examined.more » A fixed amount(0.3, 0.5, 1, and 2 mm) of MLC-leaf position deviation was simulated to each plan under following conditions: 1)only along X1 collimator; 2)with increments at both X1 and X2 collimator directions;3)with reductions at both X1 and X2 collimator directions. Dose was recalculated for each modified plans. Both original and modified plans were delivered using Turebeam STX machine and measured using both portal dosimetry and a 3D dosimeter(Delta4 of ScandiDos). Each field’s Result were compared using following three parameters: the 95% iso-dose level Conformal Index(95%CI), the spinal cord maximum dose(SCDmax), and the planned target volume(PTV) mean dose. Results: Dosimetric impacts on the 95%CI, SCDmax and the PTV mean dose are: 1)negligible if MLC-leaf position deviation only along a single collimator direction ≥1.0 mm,2)substantial if MLC-leaf position increment along both collimator directions ≥0.3 mm(95% CI decreases while SCDmax and PTV mean-dose increase), 3)substantial if MLC-leaf position reduction along both collimator directions ≥0.3 mm(95% CI first increases and then decreases while SCDmax and PTV mean-dose decrease). Different energies and doserates demonstrated comparable dosimetric impacts. Conclusion: Substantial dose deviations could happen for spinal SBRT using IMRT plan with HD-MLC if leaf position deviation ≥0.3 mm. The effects of different energy and doserate are negligible.« less
Richmond, Neil; Tulip, Rachael; Walker, Chris
2016-01-01
The aim of this work was to determine, by measurement and independent monitor unit (MU) check, the optimum method for determining collimator scatter for an Elekta Synergy linac with an Agility multileaf collimator (MLC) within Radcalc, a commercial MU calculation software package. The collimator scatter factors were measured for 13 field shapes defined by an Elekta Agility MLC on a Synergy linac with 6MV photons. The value of the collimator scatter associated with each field was also calculated according to the equation Sc=Sc(mlc)+Sc(corr)(Sc(open)-Sc(mlc)) with Sc(corr) varied between 0 and 1, where Sc(open) is the value of collimator scatter calculated from the rectangular collimator-defined field and Sc(mlc) the value using only the MLC-defined field shape by applying sector integration. From this the optimum value of the correction was determined as that which gives the minimum difference between measured and calculated Sc. Single (simple fluence modulation) and dual-arc (complex fluence modulation) treatment plans were generated on the Monaco system for prostate volumetric modulated-arc therapy (VMAT) delivery. The planned MUs were verified by absolute dose measurement in phantom and by an independent MU calculation. The MU calculations were repeated with values of Sc(corr) between 0 and 1. The values of the correction yielding the minimum MU difference between treatment planning system (TPS) and check MU were established. The empirically derived value of Sc(corr) giving the best fit to the measured collimator scatter factors was 0.49. This figure however was not found to be optimal for either the single- or dual-arc prostate VMAT plans, which required 0.80 and 0.34, respectively, to minimize the differences between the TPS and independent-check MU. Point dose measurement of the VMAT plans demonstrated that the TPS MUs were appropriate for the delivered dose. Although the value of Sc(corr) may be obtained by direct comparison of calculation with measurement, the efficacy of the value determined for VMAT-MU calculations are very much dependent on the complexity of the MLC delivery. Copyright © 2016 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Ipsen, S; Bruder, R; Schweikard, A
Purpose: While MLC tracking has been successfully used for motion compensation of moving targets, current real-time target localization methods rely on correlation models with x-ray imaging or implanted electromagnetic transponders rather than direct target visualization. In contrast, ultrasound imaging yields volumetric data in real-time (4D) without ionizing radiation. We report the first results of online 4D ultrasound-guided MLC tracking in a phantom. Methods: A real-time tracking framework was installed on a 4D ultrasound station (Vivid7 dimension, GE) and used to detect a 2mm spherical lead marker inside a water tank. The volumetric frame rate was 21.3Hz (47ms). The marker wasmore » rigidly attached to a motion stage programmed to reproduce nine tumor trajectories (five prostate, four lung). The 3D marker position from ultrasound was used for real-time MLC aperture adaption. The tracking system latency was measured and compensated by prediction for lung trajectories. To measure geometric accuracy, anterior and lateral conformal fields with 10cm circular aperture were delivered for each trajectory. The tracking error was measured as the difference between marker position and MLC aperture in continuous portal imaging. For dosimetric evaluation, 358° VMAT fields were delivered to a biplanar diode array dosimeter using the same trajectories. Dose measurements with and without MLC tracking were compared to a static reference dose using a 3%/3 mm γ-test. Results: The tracking system latency was 170ms. The mean root-mean-square tracking error was 1.01mm (0.75mm prostate, 1.33mm lung). Tracking reduced the mean γ-failure rate from 13.9% to 4.6% for prostate and from 21.8% to 0.6% for lung with high-modulation VMAT plans and from 5% (prostate) and 18% (lung) to 0% with low modulation. Conclusion: Real-time ultrasound tracking was successfully integrated with MLC tracking for the first time and showed similar accuracy and latency as other methods while holding the potential to measure target motion non-invasively. SI was supported by the Graduate School for Computing in Medicine and Life Science, German Excellence Initiative [grant DFG GSC 235/1].« less
NASA Technical Reports Server (NTRS)
Irom, Farokh; Allen, Gregory R.
2012-01-01
The space radiation environment poses a certain risk to all electronic components on Earth-orbiting and planetary mission spacecraft. In recent years, there has been increased interest in the use of high-density, commercial, nonvolatile flash memories in space because of ever-increasing data volumes and strict power requirements. They are used in a wide variety of spacecraft subsystems. At one end of the spectrum, flash memories are used to store small amounts of mission-critical data such as boot code or configuration files and, at the other end, they are used to construct multi-gigabyte data recorders that record mission science data. This report examines single-event effect (SEE) and total ionizing dose (TID) response in single-level cell (SLC) 32-Gb, multi-level cell (MLC) 64-Gb, and Triple-level (TLC) 64-Gb NAND flash memories manufactured by Micron Technology with feature size of 25 nm.
47 CFR 22.359 - Emission limitations.
Code of Federal Regulations, 2011 CFR
2011-10-01
... + 10 log (P) dB. (b) Measurement procedure. Compliance with these rules is based on the use of... contract in their station files and disclose it to prospective assignees or transferees and, upon request...
7 CFR 274.5 - Record retention and forms security.
Code of Federal Regulations, 2011 CFR
2011-01-01
... control logs, or similar controls from the point of initial receipt through the issuance and.... (2) For notices of change which initiate, update or terminate the master issuance file, the State...
ERIC Educational Resources Information Center
Tennant, Roy
1992-01-01
Explains how users can find and access information resources available on the Internet. Highlights include network information centers (NICs); lists, both formal and informal; computer networking protocols, including international standards; electronic mail; remote log-in; and file transfer. (LRW)
VizieR Online Data Catalog: CoRoT red giants abundances (Morel+, 2014)
NASA Astrophysics Data System (ADS)
Morel, T.; Miglio, A.; Lagarde, N.; Montalban, J.; Rainer, M.; Poretti, E.; Eggenberger, P.; Hekker, S.; Kallinger, T.; Mosser, B.; Valentini, M.; Carrier, F.; Hareter, M.; Mantegazza, L.
2014-02-01
The equivalent widths were measured manually assuming Gaussian profiles or Voigt profiles for the few lines with extended damping wings. Lines with an unsatisfactory fit or significantly affected by telluric features were discarded. Only values eventually retained for the analysis are provided. For the chemical abundances, the usual notation is used: [X/Y]=[log({epsilon}(X))-log({epsilon}(Y))]star - [log({epsilon}(X))-log({epsilon}(Y))]⊙ with log{epsilon}(X)=12+log[N(X)/N(H)] (N is the number density of the species). For lithium, the following notation is used: [Li/H]=log(N(Li))star-log(N(Li))⊙. The adopted solar abundances are taken from Grevesse & Sauval (1998SSRv...85..161G), except for Li for which we adopt our derived values: log({epsilon}(Li))⊙=1.09 and 1.13 in LTE and NLTE, respectively (see text). All the abundances are computed under the assumption of LTE, except Li for which values corrected for departures from LTE using the data of Lind et al. (2009A&A...503..541L) are also provided. All the quoted error bars are 1-sigma uncertainties. (6 data files).
USDA-ARS?s Scientific Manuscript database
Myosin light chain (MLC) plays a vital role in cell and muscle functions and has been identified as an allergen in close species. In this study, MLC with the molecular mass of 18kDa was purified from crayfish (Procambarus clarkii) muscle fibrils. Its physicochemical characterization showed that the...
VizieR Online Data Catalog: NGC 2264, NGC 2547 and NGC 2516 stellar radii (Jackson+, 2016)
NASA Astrophysics Data System (ADS)
Jackson, R. J.; Jeffries, R. D.; Randich, S.; Bragaglia, A.; Carraro, G.; Costado, M. T.; Flaccomio, E.; Lanzafame; Lardo, C.; Monaco, L.; Morbidelli, L.; Smiljanic, R.; Zaggia, S.
2015-11-01
File Table1.dat contains Photometric and spectroscopic data of GES Survey targets in clusters in NGC 2547, NGC 2516, NGC 22264 downloaded from the Edinburugh GES archive (http://ges/roe.ac.uk/) . Photometric data comprised the (Cousins) I magnitude and 2MASS J, H and K magnitudes. Spectroscopic data comprises the signal to noise ratio, S/N of the target spectrum, the radial velocity, RV (in km/s), the projected equatorial velocity, vsini (in km/s), the number of separate observations co-added to produce the target spectrum and the log of effective temperature (logTeff) of the template spectrum fitted to measure RV and vsini. The absolute precision in RV, pRV (in km/s) and relative precision vsini (pvsini) were estimated, as a function of the logTeff, vsini and S/N, using the prescription described in Jackson et al. (2015A&A...580A..75J, Cat. J/A+A/580/A75). File Table3.dat contains measured and calculated properties of cluster targets with resolved vsini and a reported rotation period. The cluster name, right ascension, RA (deg) and declination, Dec (deg) are given for targets with measured periods given in the literature. Dynamic properties comprise: the radial velocity, RV (in km/s), the absolute precision in RV, pRV (km/s), the projected equatorial velocity, vsini (in km/s), the relative precision in vsini (pvsini) and the rotational period (in days). Also shown are values of absolute K magnitude, MK log of luminosity, log L (in solar units) and probability of cluster membership estimated using cluster data given in the text. Period shows reported values of cluster taken from the literature Estimated values of the projected radius, Rsini (in Rsolar) and uncertainty in projected radius, e_Rsini (in Rsolar) are given for targets where vsini>5km/s and pvsini>0.2. The final column shows a flag which is set to 1 for targets in cluster NGC 2264 where a (H-K) versus (J-H) colour-colour plot indicates possible infra-red excess. Period shows reported values of cluster taken from the literature (2 data files).
Optimizing Earth Data Search Ranking using Deep Learning and Real-time User Behaviour
NASA Astrophysics Data System (ADS)
Jiang, Y.; Yang, C. P.; Armstrong, E. M.; Huang, T.; Moroni, D. F.; McGibbney, L. J.; Greguska, F. R., III
2017-12-01
Finding Earth science data has been a challenging problem given both the quantity of data available and the heterogeneity of the data across a wide variety of domains. Current search engines in most geospatial data portals tend to induce end users to focus on one single data characteristic dimension (e.g., term frequency-inverse document frequency (TF-IDF) score, popularity, release date, etc.). This approach largely fails to take account of users' multidimensional preferences for geospatial data, and hence may likely result in a less than optimal user experience in discovering the most applicable dataset out of a vast range of available datasets. With users interacting with search engines, sufficient information is already hidden in the log files. Compared with explicit feedback data, information that can be derived/extracted from log files is virtually free and substantially more timely. In this dissertation, I propose an online deep learning framework that can quickly update the learning function based on real-time user clickstream data. The contributions of this framework include 1) a log processor that can ingest, process and create training data from web logs in a real-time manner; 2) a query understanding module to better interpret users' search intent using web log processing results and metadata; 3) a feature extractor that identifies ranking features representing users' multidimensional interests of geospatial data; and 4) a deep learning based ranking algorithm that can be trained incrementally using user behavior data. The search ranking results will be evaluated using precision at K and normalized discounted cumulative gain (NDCG).
43 CFR 2743.3 - Leased disposal sites.
Code of Federal Regulations, 2011 CFR
2011-10-01
... review of all records and inspection reports on file with the Bureau of Land Management, State, and local... landfill concerning site management and a review of all reports and logs pertaining to the type and amount...
25 CFR 214.13 - Diligence; annual expenditures; mining records.
Code of Federal Regulations, 2011 CFR
2011-04-01
... within 90 days after an ore body of sufficient quantity is discovered, and shown by the logs or records.... Lessee shall, before commencing operations, file with the superintendent a plat and preliminary statement...
47 CFR 22.861 - Emission limitations.
Code of Federal Regulations, 2011 CFR
2011-10-01
... below the transmitting power (P) by a factor of at least 43 + 10 log (P) dB. (b) Measurement procedure... maintain a copy of the contract in their station files and disclose it to prospective assignees or...
Development of Cross-Platform Software for Well Logging Data Visualization
NASA Astrophysics Data System (ADS)
Akhmadulin, R. K.; Miraev, A. I.
2017-07-01
Well logging data processing is one of the main sources of information in the oil-gas field analysis and is of great importance in the process of its development and operation. Therefore, it is important to have the software which would accurately and clearly provide the user with processed data in the form of well logs. In this work, there have been developed a software product which not only has the basic functionality for this task (loading data from .las files, well log curves display, etc.), but can be run in different operating systems and on different devices. In the article a subject field analysis and task formulation have been performed, and the software design stage has been considered. At the end of the work the resulting software product interface has been described.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Yang, B; Wong, R; Lam, W
Purpose: To develop a practical method for routine QA of the MLC of a Tomotherapy unit using ArcCheck. Methods: Two standard test plans were used in this study. One was a helical test, in which the central leaves No. 32 and 33 opened simultaneously for 277.8ms at projections centered at 0°, 120° and 240° gantry angles. The other test plan was a static test with the gantry angle set at 0°, 45°, 90° and 135° respectively and leaves No. 32 and 33 opened sequentially for total 20s which was further divided into eleven or ten segments at each beam angle.more » The ArcCheck was isocentrically set up and adjusted for couch sag. Movie files which took a snapshot exposure every 50ms were recorded. The start/stop time of leaf open was decided by the ramp-up/ramp-down of the detectors. Results: The percentage differences between measured and planned leaf open time were calculated to be within 0.5% in all the tests. In static test, if leaves are synchronized perfectly, the sum of the two detectors’ signals after normalization should equal one when the leaves are in transition. Our results showed mean values of 0.982, 0.983, 0.978 and 0.995 at static gantry angle 0°, 45°, 90° and 135° respectively. Conclusion: A method for estimating the Tomotherapy binary MLC leaf open time using ArcCheck is proposed and proved to be precise enough to verify the planned leaf open time as small as 277.8ms. This method also makes the observation and quantification of the synchronization of leaves possible.« less
Yamauchi, Kaori; Li, Junjun; Morikawa, Kumi; Liu, Li; Shirayoshi, Yasuaki; Nakatsuji, Norio; Elliott, David A; Hisatome, Ichiro; Suemori, Hirofumi
2018-01-01
Human pluripotent stem cell (hPSC)-derived cardiomyocytes (CMs) are a promising source for cell transplantation into the damaged heart, which has limited regenerative ability. Many methods have been developed to obtain large amounts of functional CMs from hPSCs for therapeutic applications. However, during the differentiation process, a mixed population of various cardiac cells, including ventricular, atrial, and pacemaker cells, is generated, which hampers the proper functional analysis and evaluation of cell properties. Here, we established NKX2-5 eGFP/w and MLC2v mCherry/w hPSC double knock-ins that allow for labeling, tracing, purification, and analysis of the development of ventricular cells from early to late stages. As with the endogenous transcriptional activities of these genes, MLC2v-mCherry expression following NKX2-5-eGFP expression was observed under previously established culture conditions, which mimic the in vivo cardiac developmental process. Patch-clamp and microelectrode array electrophysiological analyses showed that the NKX2-5 and MLC2v double-positive cells possess ventricular-like properties. The results demonstrate that the NKX2-5 eGFP/w and MLC2v mCherry/w hPSCs provide a powerful model system to capture region-specific cardiac differentiation from early to late stages. Our study would facilitate subtype-specific cardiac development and functional analysis using the hPSC-derived sources. Copyright © 2017 Elsevier Inc. All rights reserved.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Nguyen, N; Knutson, N; Schmidt, M
Purpose: To verify a method used to automatically acquire jaw, MLC, collimator and couch star shots for a Varian TrueBeam linear accelerator utilizing Developer Mode and an Electronic Portal Imaging Device (EPID). Methods: An XML script was written to automate motion of the jaws, MLC, collimator and couch in TrueBeam Developer Mode (TBDM) to acquire star shot measurements. The XML script also dictates MV imaging parameters to facilitate automatic acquisition and recording of integrated EPID images. Since couch star shot measurements cannot be acquired using a combination of EPID and jaw/MLC collimation alone due to a fixed imager geometry, amore » method utilizing a 5mm wide steel ruler placed on the table and centered within a 15×15cm2 open field to produce a surrogate of the narrow field aperture was investigated. Four individual star shot measurements (X jaw, Y jaw, MLC and couch) were obtained using our proposed as well as traditional film-based method. Integrated EPID images and scanned measurement films were analyzed and compared. Results: Star shot (X jaw, Y jaw, MLC and couch) measurements were obtained in a single 5 minute delivery using the TBDM XML script method compared to 60 minutes for equivalent traditional film measurements. Analysis of the images and films demonstrated comparable isocentricity results, agreeing within 0.3mm of each other. Conclusion: The presented automatic approach of acquiring star shot measurements using TBDM and EPID has proven to be more efficient than the traditional film approach with equivalent results.« less
Real-time auto-adaptive margin generation for MLC-tracked radiotherapy
NASA Astrophysics Data System (ADS)
Glitzner, M.; Fast, M. F.; de Senneville, B. Denis; Nill, S.; Oelfke, U.; Lagendijk, J. J. W.; Raaymakers, B. W.; Crijns, S. P. M.
2017-01-01
In radiotherapy, abdominal and thoracic sites are candidates for performing motion tracking. With real-time control it is possible to adjust the multileaf collimator (MLC) position to the target position. However, positions are not perfectly matched and position errors arise from system delays and complicated response of the electromechanic MLC system. Although, it is possible to compensate parts of these errors by using predictors, residual errors remain and need to be compensated to retain target coverage. This work presents a method to statistically describe tracking errors and to automatically derive a patient-specific, per-segment margin to compensate the arising underdosage on-line, i.e. during plan delivery. The statistics of the geometric error between intended and actual machine position are derived using kernel density estimators. Subsequently a margin is calculated on-line according to a selected coverage parameter, which determines the amount of accepted underdosage. The margin is then applied onto the actual segment to accommodate the positioning errors in the enlarged segment. The proof-of-concept was tested in an on-line tracking experiment and showed the ability to recover underdosages for two test cases, increasing {{V}90 %} in the underdosed area about 47 % and 41 % , respectively. The used dose model was able to predict the loss of dose due to tracking errors and could be used to infer the necessary margins. The implementation had a running time of 23 ms which is compatible with real-time requirements of MLC tracking systems. The auto-adaptivity to machine and patient characteristics makes the technique a generic yet intuitive candidate to avoid underdosages due to MLC tracking errors.
HRR Upgrade to mass loss calorimeter and modified Schlyter test for FR Wood
Mark A. Dietenberger; Charles R. Boardman
2013-01-01
Enhanced Heat Release Rate (HRR) methodology has been extended to the Mass Loss Calorimeter (MLC) and the Modified Schlyter flame spread test to evaluate fire retardant effectiveness used on wood based materials. Modifications to MLC include installation of thermopile on the chimney walls to correct systematic errors to the sensible HRR calculations to account for...
NASA Technical Reports Server (NTRS)
Brieda, Lubos
2015-01-01
This talk presents 3 different tools developed recently for contamination analysis:HTML QCM analyzer: runs in a web browser, and allows for data analysis of QCM log filesJava RGA extractor: can load in multiple SRS.ana files and extract pressure vs. time dataC++ Contamination Simulation code: 3D particle tracing code for modeling transport of dust particulates and molecules. Uses residence time to determine if molecules stick. Particulates can be sampled from IEST-STD-1246 and be accelerated by aerodynamic forces.
AliEn—ALICE environment on the GRID
NASA Astrophysics Data System (ADS)
Saiz, P.; Aphecetche, L.; Bunčić, P.; Piskač, R.; Revsbech, J.-E.; Šego, V.; Alice Collaboration
2003-04-01
AliEn ( http://alien.cern.ch) (ALICE Environment) is a Grid framework built on top of the latest Internet standards for information exchange and authentication (SOAP, PKI) and common Open Source components. AliEn provides a virtual file catalogue that allows transparent access to distributed datasets and a number of collaborating Web services which implement the authentication, job execution, file transport, performance monitor and event logging. In the paper we will present the architecture and components of the system.
Parallel compression of data chunks of a shared data object using a log-structured file system
DOE Office of Scientific and Technical Information (OSTI.GOV)
Bent, John M.; Faibish, Sorin; Grider, Gary
2016-10-25
Techniques are provided for parallel compression of data chunks being written to a shared object. A client executing on a compute node or a burst buffer node in a parallel computing system stores a data chunk generated by the parallel computing system to a shared data object on a storage node by compressing the data chunk; and providing the data compressed data chunk to the storage node that stores the shared object. The client and storage node may employ Log-Structured File techniques. The compressed data chunk can be de-compressed by the client when the data chunk is read. A storagemore » node stores a data chunk as part of a shared object by receiving a compressed version of the data chunk from a compute node; and storing the compressed version of the data chunk to the shared data object on the storage node.« less
Ambrozy, C; Kolar, N A; Rattay, F
2010-01-01
For measurement value logging of board angle values during balance training, it is necessary to develop a measurement system. This study will provide data for a balance study using the smartcard. The data acquisition comes automatically. An individually training plan for each proband is necessary. To store the proband identification a smartcard with an I2C data bus protocol and an E2PROM memory system is used. For reading the smartcard data a smartcard reader is connected via universal serial bus (USB) to a notebook. The data acquisition and smartcard read programme is designed with Microsoft® Visual C#. A training plan file contains the individual training plan for each proband. The data of the test persons are saved in a proband directory. Each event is automatically saved as a log-file for the exact documentation. This system makes study development easy and time-saving.
VizieR Online Data Catalog: GOALS sample PACS and SPIRE fluxes (Chu+, 2017)
NASA Astrophysics Data System (ADS)
Chu, J. K.; Sanders, D. B.; Larson, K. L.; Mazzarella, J. M.; Howell, J. H.; Diaz-Santos, T.; Xu, K. C.; Paladini, R.; Schulz, B.; Shupe, D.; Appleton, P.; Armus, L.; Billot, N.; Chan, B. H. P.; Evans, A. S.; Fadda, D.; Frayer, D. T.; Haan, S.; Ishida, C. M.; Iwasawa, K.; Kim, D.-C.; Lord, S.; Murphy, E.; Petric, A.; Privon, G. C.; Surace, J. A.; Treister, E.
2017-06-01
The IRAS RBGS contains 179 LIRGs (log(LIR/Lȯ)= 22 ultra-luminous infrared galaxies (ULIRGs: log(LIR/Lȯ)>=12.0); these 201 total objects comprise the GOALS sample (Armus et al. 2009), a statistically complete flux-limited sample of infrared-luminous galaxies in the local universe. This paper presents imaging and photometry for all 201 LIRGs and LIRG systems in the IRAS RBGS that were observed during our GOALS Herschel OT1 program. (4 data files).
Rowe, Steven P; Siddiqui, Adeel; Bonekamp, David
2014-07-01
To create novel radiology key image software that is easy to use for novice users, incorporates elements adapted from social networking Web sites, facilitates resident and fellow education, and can serve as the engine for departmental sharing of interesting cases and follow-up studies. Using open-source programming languages and software, radiology key image software (the key image and case log application, KICLA) was developed. This system uses a lightweight interface with the institutional picture archiving and communications systems and enables the storage of key images, image series, and cine clips. It was designed to operate with minimal disruption to the radiologists' daily workflow. Many features of the user interface have been inspired by social networking Web sites, including image organization into private or public folders, flexible sharing with other users, and integration of departmental teaching files into the system. We also review the performance, usage, and acceptance of this novel system. KICLA was implemented at our institution and achieved widespread popularity among radiologists. A large number of key images have been transmitted to the system since it became available. After this early experience period, the most commonly encountered radiologic modalities are represented. A survey distributed to users revealed that most of the respondents found the system easy to use (89%) and fast at allowing them to record interesting cases (100%). Hundred percent of respondents also stated that they would recommend a system such as KICLA to their colleagues. The system described herein represents a significant upgrade to the Digital Imaging and Communications in Medicine teaching file paradigm with efforts made to maximize its ease of use and inclusion of characteristics inspired by social networking Web sites that allow the system additional functionality such as individual case logging. Copyright © 2014 AUR. Published by Elsevier Inc. All rights reserved.
Distributed Storage Algorithm for Geospatial Image Data Based on Data Access Patterns.
Pan, Shaoming; Li, Yongkai; Xu, Zhengquan; Chong, Yanwen
2015-01-01
Declustering techniques are widely used in distributed environments to reduce query response time through parallel I/O by splitting large files into several small blocks and then distributing those blocks among multiple storage nodes. Unfortunately, however, many small geospatial image data files cannot be further split for distributed storage. In this paper, we propose a complete theoretical system for the distributed storage of small geospatial image data files based on mining the access patterns of geospatial image data using their historical access log information. First, an algorithm is developed to construct an access correlation matrix based on the analysis of the log information, which reveals the patterns of access to the geospatial image data. Then, a practical heuristic algorithm is developed to determine a reasonable solution based on the access correlation matrix. Finally, a number of comparative experiments are presented, demonstrating that our algorithm displays a higher total parallel access probability than those of other algorithms by approximately 10-15% and that the performance can be further improved by more than 20% by simultaneously applying a copy storage strategy. These experiments show that the algorithm can be applied in distributed environments to help realize parallel I/O and thereby improve system performance.
Shariati, Gholamreza; Hamid, Mohammad; Saberi, Alihossein; Andashti, Behnaz; Galehdari, Hamid
2015-02-01
Megalencephalic leukoencephalopathy (MLC) is a rare neurological disorder with an autosomal recessive pattern. Clinical diagnosis was based on macrocephaly, recurrent seizure, and magnetic resonance imaging (MRI). Here we report first finding of a novel homozygous single base deletion in the MLC1 gene in an affected Iranian child causing a premature stop codon (p.L150fs.160X).
Development of a Multileaf Collimator for Proton Radiotherapy
2007-06-01
for proton radiotherapy, and the first year of the project to develop image guided treatment protocols for proton therapy . This research...multileaf collimator (MLC) for proton therapy and investigates the issues that must be resolved to use an MLC in proton therapy . The second technology...the contract included three development agreements directly related to the work supported by this grant to develop technology for proton therapy .
ERIC Educational Resources Information Center
Viner, Jane; Lucas, Amanda; Ricchini, Tracey; Ri, Regina
2010-01-01
This workshop paper explores the Web 2.0 journey of the MLC Libraries' teacher-librarians, librarian, library and audio visual technicians. Our journey was initially inspired by Will Richardson and supported by the School Library Association of Victoria (SLAV) Web 2.0 professional development program. The 12 week technological skills program…
DOE Office of Scientific and Technical Information (OSTI.GOV)
Pogson, EM; Liverpool and Macarthur Cancer Therapy Centres, Liverpool, NSW; Ingham Institute for Applied Medical Research, Sydney, NSW
Purpose: To identify the robustness of different treatment techniques in respect to simulated linac errors on the dose distribution to the target volume and organs at risk for step and shoot IMRT (ssIMRT), VMAT and Autoplan generated VMAT nasopharynx plans. Methods: A nasopharynx patient dataset was retrospectively replanned with three different techniques: 7 beam ssIMRT, one arc manual generated VMAT and one arc automatically generated VMAT. Treatment simulated uncertainties: gantry, collimator, MLC field size and MLC shifts, were introduced into these plans at increments of 5,2,1,−1,−2 and −5 (degrees or mm) and recalculated in Pinnacle. The mean and maximum dosesmore » were calculated for the high dose PTV, parotids, brainstem, and spinal cord and then compared to the original baseline plan. Results: Simulated gantry angle errors have <1% effect on the PTV, ssIMRT is most sensitive. The small collimator errors (±1 and ±2 degrees) impacted the mean PTV dose by <2% for all techniques, however for the ±5 degree errors mean target varied by up to 7% for the Autoplan VMAT and 10% for the max dose to the spinal cord and brain stem, seen in all techniques. The simulated MLC shifts introduced the largest errors for the Autoplan VMAT, with the larger MLC modulation presumably being the cause. The most critical error observed, was the MLC field size error, where even small errors of 1 mm, caused significant changes to both the PTV and the OAR. The ssIMRT is the least sensitive and the Autoplan the most sensitive, with target errors of up to 20% over and under dosages observed. Conclusion: For a nasopharynx patient the plan robustness observed is highest for the ssIMRT plan and lowest for the Autoplan generated VMAT plan. This could be caused by the more complex MLC modulation seen for the VMAT plans. This project is supported by a grant from NSW Cancer Council.« less
Qiu, Jianguo; Chen, Shuting; Wu, Hong
2014-05-01
Hepatic resection (HRE) combined with radiofrequency ablation (RFA) offers a surgical option to a group of patients with multiple and bilobar liver malignancies who are traditionally unresectable for inadequate functional hepatic reserve. The aims of the present study were to assess the perioperative outcomes, recurrence, and long-term survival rates for patients treated with HRE plus RFA in the management of primary hepatocellular carcinoma (HCC) and metastatic liver cancer (MLC). Data from all consecutive patients with primary and secondary hepatic malignancies who were treated with HRE combined with RFA between 2007 and 2013 were prospectively collected and retrospectively reviewed. A total of 112 patients, with 368 hepatic tumors underwent HRE combined with ultrasound-guided RFA, were included in the present study. There were 40 cases of HCC with 117 tumors and 72 cases of MLC with 251 metastases. Most cases of liver metastases originated from the gastrointestinal tract (44, 61.1%). Other uncommon lesions included breast cancer (5, 6.9%), pancreatic cancer (3, 4.2%), lung cancer (4, 5.6%), cholangiocarcinoma (4, 5.6%), and so on. The ablation success rates were 93.3% for HCC and 96.7% for MLC. The 1-, 2-, 3-, 4-, and 5-y overall recurrence rates were 52.5%, 59.5%, 72.3%, 75%, and 80% for the HCC group and 44.4%, 52.7%, 56.1%, 69.4%, and 77.8% for the MLC group, respectively. The 1-, 2-, 3-, 4-, and 5-y overall survival rates for the HCC patients were 67.5%, 50%, 32.5%, 22.5%, and 12.5% and for the MLC patients were 66.5%, 55.5%, 50%, 30.5%, and 19.4%, respectively. The corresponding recurrence-free survival rates for the HCC patients were 52.5%, 35%, 22.5%, 15%, and 10% and for the MLC patients were 58.3%, 41.6%, 23.6%, 16.9%, and 12.5%, respectively. HRE combined with RFA provides an effective treatment approach for patients with primary and secondary liver malignancies who are initially unsuitable for radical resection, with high local tumor control rates and promising survival data. Copyright © 2014 Elsevier Inc. All rights reserved.
TomoTherapy MLC verification using exit detector data
DOE Office of Scientific and Technical Information (OSTI.GOV)
Chen Quan; Westerly, David; Fang Zhenyu
2012-01-15
Purpose: Treatment delivery verification (DV) is important in the field of intensity modulated radiation therapy (IMRT). While IMRT and image guided radiation therapy (IGRT), allow us to create more conformal plans and enables the use of tighter margins, an erroneously executed plan can have detrimental effects on the treatment outcome. The purpose of this study is to develop a DV technique to verify TomoTherapy's multileaf collimator (MLC) using the onboard mega-voltage CT detectors. Methods: The proposed DV method uses temporal changes in the MVCT detector signal to predict actual leaf open times delivered on the treatment machine. Penumbra and scatteredmore » radiation effects may produce confounding results when determining leaf open times from the raw detector data. To reduce the impact of the effects, an iterative, Richardson-Lucy (R-L) deconvolution algorithm is applied. Optical sensors installed on each MLC leaf are used to verify the accuracy of the DV technique. The robustness of the DV technique is examined by introducing different attenuation materials in the beam. Additionally, the DV technique has been used to investigate several clinical plans which failed to pass delivery quality assurance (DQA) and was successful in identifying MLC timing discrepancies as the root cause. Results: The leaf open time extracted from the exit detector showed good agreement with the optical sensors under a variety of conditions. Detector-measured leaf open times agreed with optical sensor data to within 0.2 ms, and 99% of the results agreed within 8.5 ms. These results changed little when attenuation was added in the beam. For the clinical plans failing DQA, the dose calculated from reconstructed leaf open times played an instrumental role in discovering the root-cause of the problem. Throughout the retrospective study, it is found that the reconstructed dose always agrees with measured doses to within 1%. Conclusions: The exit detectors in the TomoTherapy treatment systems can provide valuable information about MLC behavior during delivery. A technique to estimate the TomoTherapy binary MLC leaf open time from exit detector signals is described. This technique is shown to be both robust and accurate for delivery verification.« less
Rosca, Florin; Lorenz, Friedlieb; Hacker, Fred L; Chin, Lee M; Ramakrishna, Naren; Zygmanski, Piotr
2006-06-01
We have designed and implemented a new stereotactic linac QA test with stereotactic precision. The test is used to characterize gantry sag, couch wobble, cone placement, MLC offsets, and room lasers' positions relative to the radiation isocenter. Two MLC star patterns, a cone pattern, and the laser line patterns are recorded on the same imaging medium. Phosphor plates are used as imaging medium due to their sensitivity to red light. The red light of room lasers erases some of the irradiation information stored on the phosphor plates enabling accurate and direct measurements for the position of room lasers and radiation isocenter. Using film instead of the phosphor plate as imaging medium is possible, however, it is less practical. The QA method consists of irradiating four phosphor plates that record the gantry sag between the 0 degrees and 180 degrees gantry angles, the position and stability of couch rotational axis, the sag between the 90 degrees and 270 degrees gantry angles, the accuracy of cone placement on the collimator, the MLC offsets from the collimator rotational axis, and the position of laser lines relative to the radiation isocenter. The estimated accuracy of the method is +/- 0.2 mm. The observed reproducibility of the method is about +/- 0.1 mm. The total irradiation/ illumination time is about 10 min per image. Data analysis, including the phosphor plate scanning, takes less than 5 min for each image. The method characterizes the radiation isocenter geometry with the high accuracy required for the stereotactic radiosurgery. In this respect, it is similar to the standard ball test for stereotactic machines. However, due to the usage of the MLC instead of the cross-hair/ball, it does not depend on the cross-hair/ball placement errors with respect to the lasers and it provides more information on the mechanical integrity of the linac/couch/laser system. Alternatively, it can be used as a highly accurate QA procedure for the nonstereotactic machines. Noteworthy is its ability to characterize the MLC position accuracy, which is an important factor in IMRT delivery.
Dynamic tumor tracking using the Elekta Agility MLC
DOE Office of Scientific and Technical Information (OSTI.GOV)
Fast, Martin F., E-mail: martin.fast@icr.ac.uk; Nill, Simeon, E-mail: simeon.nill@icr.ac.uk; Bedford, James L.
2014-11-01
Purpose: To evaluate the performance of the Elekta Agility multileaf collimator (MLC) for dynamic real-time tumor tracking. Methods: The authors have developed a new control software which interfaces to the Agility MLC to dynamically program the movement of individual leaves, the dynamic leaf guides (DLGs), and the Y collimators (“jaws”) based on the actual target trajectory. A motion platform was used to perform dynamic tracking experiments with sinusoidal trajectories. The actual target positions reported by the motion platform at 20, 30, or 40 Hz were used as shift vectors for the MLC in beams-eye-view. The system latency of the MLCmore » (i.e., the average latency comprising target device reporting latencies and MLC adjustment latency) and the geometric tracking accuracy were extracted from a sequence of MV portal images acquired during irradiation for the following treatment scenarios: leaf-only motion, jaw + leaf motion, and DLG + leaf motion. Results: The portal imager measurements indicated a clear dependence of the system latency on the target position reporting frequency. Deducting the effect of the target frequency, the leaf adjustment latency was measured to be 38 ± 3 ms for a maximum target speed v of 13 mm/s. The jaw + leaf adjustment latency was 53 ± 3 at a similar speed. The system latency at a target position frequency of 30 Hz was in the range of 56–61 ms for the leaves (v ≤ 31 mm/s), 71–78 ms for the jaw + leaf motion (v ≤ 25 mm/s), and 58–72 ms for the DLG + leaf motion (v ≤ 59 mm/s). The tracking accuracy showed a similar dependency on the target position frequency and the maximum target speed. For the leaves, the root-mean-squared error (RMSE) was between 0.6–1.5 mm depending on the maximum target speed. For the jaw + leaf (DLG + leaf) motion, the RMSE was between 0.7–1.5 mm (1.9–3.4 mm). Conclusions: The authors have measured the latency and geometric accuracy of the Agility MLC, facilitating its future use for clinical tracking applications.« less
47 CFR 22.917 - Emission limitations for cellular equipment.
Code of Federal Regulations, 2011 CFR
2011-10-01
... frequency ranges must be attenuated below the transmitting power (P) by a factor of at least 43 + 10 log(P... such contract shall maintain a copy of the contract in their station files and disclose it to...
Shende, Ravindra; Patel, Ganesh
2017-01-01
Objective of present study is to determine optimum value of DLG and its validation prior to being incorporated in TPS for Varian TrueBeam™ millennium 120 leaves MLC. Partial transmission through the rounded leaf ends of the Multi Leaf Collimator (MLC) causes a conflict between the edges of the light field and radiation field. Parameter account for this partial transmission is called Dosimetric Leaf Gap (DLG). The complex high precession technique, such as Intensity Modulated Radiation Therapy (IMRT), entails the modeling of optimum value of DLG inside Eclipse Treatment Planning System (TPS) for precise dose calculation. Distinct synchronized uniformed extension of sweeping dynamic MLC leaf gap fields created by Varian MLC shaper software were use to determine DLG. DLG measurements performed with both 0.13 cc semi-flex ionization chamber and 2D-Array I-Matrix were used to validate the DLG; similarly, values of DLG from TPS were estimated from predicted dose. Similar mathematical approaches were employed to determine DLG from delivered and TPS predicted dose. DLG determined from delivered dose measured with both ionization chamber (DLG Ion ) and I-Matrix (DLG I-Matrix ) compared with DLG estimate from TPS predicted dose (DLG TPS ). Measurements were carried out for all available 6MV, 10MV, 15MV, 6MVFFF and 10MVFFF beam energies. Maximum and minimum DLG deviation between measured and TPS calculated DLG was found to be 0.2 mm and 0.1 mm, respectively. Both of the measured DLGs (DLG Ion and DLG I-Matrix ) were found to be in a very good agreement with estimated DLG from TPS (DLG TPS ). Proposed method proved to be helpful in verifying and validating the DLG value prior to its clinical implementation in TPS.
Stępnik, Katarzyna E; Malinowska, Irena; Maciejewska, Małgorzata
2016-06-01
The determination of free drug concentration is a very important issue in the field of pharmacology because only the unbound drug fraction can achieve a pharmacological effect. Due to the ability to solubilize many different compounds in micellar aggregates, micellar liquid chromatography (MLC) can be used for direct determination of free drug concentration. Proteins are not retained on the stationary phase probably due to the formation of protein - surfactant complexes which are excluded from the pores of stationary phase. The micellar method is simple and fast. It does not require any pre-preparation of the tested samples for analysis. The main aim of this paper is to demonstrate a completely new applicability of the analytical use of MLC concerning the determination of free drug concentration in the standard solution of human serum albumin. The well-known adsorption method using RP-HPLC and the spectrophotometric technique was applied as the reference method. The results show that the free drug concentration value obtained in the MLC system (based on the RP-8 stationary phase and CTAB) is similar to that obtained by the adsorption method: both RP-HPLC (95.83μgmL(-1), 79.86% of free form) and spectrophotometry (95.71μgmL(-1), 79.76%). In the MLC the free drug concentration was 93.98μgmL(-1) (78.3%). This indicates that the obtained results are within the analytical range of % of free ampicillin fraction and the MLC with direct sample injection can be treated like a promising method for the determination of free drug concentration. Copyright © 2016 Elsevier B.V. All rights reserved.
Nakayama, Shinichi; Monzen, Hajime; Oonishi, Yuuichi; Mizote, Rika; Iramina, Hiraku; Kaneshige, Souichirou; Mizowaki, Takashi
2015-06-01
Photographic film is widely used for the dose distribution verification of intensity-modulated radiation therapy (IMRT). However, analysis for verification of the results is subjective. We present a novel method for marking the isocenter using irradiation from a megavoltage (MV) beam transmitted through slits in a multi-leaf collimator (MLC). We evaluated the effect of the marking irradiation at 500 monitor units (MU) on the total transmission through the MLC using an ionization chamber and Radiochromic Film. Film dosimetry was performed for quality assurance (QA) of IMRT plans. Three methods of registration were used for each film: marking by irradiating with an MV beam through slits in the MLC (MLC-IC); marking with a fabricated phantom (Phantom-IC); and a subjective method based on isodose lines (Manual). Each method was subjected to local γ-analysis. The effect of the marking irradiation on the total transmission was 0.16%, as measured by a ionization chamber at a 10-cm depth in a solid phantom, while the inter-leaf transmission was 0.3%, determined from the film. The mean pass rates for each registration method agreed within ± 1% when the criteria used were a distance-to-agreement (DTA) of 3 mm and a dose difference (DD) of 3%. For DTA/DD criteria of 2mm/3%, the pass rates in the sagittal plane were 96.09 ± 0.631% (MLC-IC), 96.27 ± 0.399% (Phantom-IC), and 95.62 ± 0.988% (Manual). The present method is a versatile and useful method of improving the objectivity of film dosimetry for IMRT QA. Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
TU-FG-201-04: Computer Vision in Autonomous Quality Assurance of Linear Accelerators
DOE Office of Scientific and Technical Information (OSTI.GOV)
Yu, H; Jenkins, C; Yu, S
Purpose: Routine quality assurance (QA) of linear accelerators represents a critical and costly element of a radiation oncology center. Recently, a system was developed to autonomously perform routine quality assurance on linear accelerators. The purpose of this work is to extend this system and contribute computer vision techniques for obtaining quantitative measurements for a monthly multi-leaf collimator (MLC) QA test specified by TG-142, namely leaf position accuracy, and demonstrate extensibility for additional routines. Methods: Grayscale images of a picket fence delivery on a radioluminescent phosphor coated phantom are captured using a CMOS camera. Collected images are processed to correct formore » camera distortions, rotation and alignment, reduce noise, and enhance contrast. The location of each MLC leaf is determined through logistic fitting and a priori modeling based on knowledge of the delivered beams. Using the data collected and the criteria from TG-142, a decision is made on whether or not the leaf position accuracy of the MLC passes or fails. Results: The locations of all MLC leaf edges are found for three different picket fence images in a picket fence routine to 0.1mm/1pixel precision. The program to correct for image alignment and determination of leaf positions requires a runtime of 21– 25 seconds for a single picket, and 44 – 46 seconds for a group of three pickets on a standard workstation CPU, 2.2 GHz Intel Core i7. Conclusion: MLC leaf edges were successfully found using techniques in computer vision. With the addition of computer vision techniques to the previously described autonomous QA system, the system is able to quickly perform complete QA routines with minimal human contribution.« less
Massunari, Loiane; Novais, Renata Zoccal; Oliveira, Márcio Teixeira; Valentim, Diego; Dezan Junior, Eloi; Duque, Cristiane
2017-01-01
Psidium cattleianum (PC) has been displaying inhibitory effect against a variety of microorganisms, but this effect has not yet been tested against endodontic pathogens. The aim of this study was to evaluate the antimicrobial activity and biocompatibility of the aqueous (PCAE) and hydroethanolic (PCHE) extracts from Psidium cattleianum (PC) leaves. Minimum inhibitory concentration (MIC) and minimum lethal concentration (MLC) were determined using the microdilution broth method in order to analyze the antimicrobial effect against Enterococcus faecalis, Pseudomonas aeruginosa, Actinomyces israelii and Candida albicans in planktonic conditions. Biofilm assays were conducted only with the extracts that were able to determine the MLC for microorganisms in planktonic conditions. Immediate and late tissue reactions against PC extracts were evaluated using edemogenic test and histological analysis of subcutaneous implants in Wistar rats. The results showed that the MIC and MLC values ranged between 0.25 and 4 mg/mL. The MLC obtained for PCHE inhibited 100% growth of all the tested strains, except for C. albicans. PCAE had the same effect for E. faecalis and P. aeruginosa. Both PC extracts were able to eliminate E. faecalis biofilms and only the PCHE eliminated P. aeruginosa biofilms. The positive controls inhibited the growth of all tested strains in MIC and MLC essays, but no CHX tested concentrations were able to eliminate A. israelii biofilm. PCAE caused a discrete increase in the edema over time, while PCHE caused a higher initial edema, which decreased progressively. Both PCAE and PCHE extracts were biocompatible, but PCHE showed better results with slight levels of inflammation at 28 days. In conclusion, PCHE was biocompatible and presented better antimicrobial effect against important pathogens associated with persistent endodontic infections.
NASA Astrophysics Data System (ADS)
Langhans, Marco; Echner, Gernot; Runz, Armin; Baumann, Martin; Xu, Mark; Ueltzhöffer, Stefan; Häring, Peter; Schlegel, Wolfgang
2015-04-01
According to the Directory of Radiotherapy Centres (DIRAC) there are 2348 Cobalt-60 (Co-60) teletherapy units worldwide, most of them in low and middle income countries, compared to 11046 clinical accelerators. To improve teletherapy with Co-60, a mechanical Multi-Leaf Collimator (MLC) was developed, working with pneumatic pressure and thus independent of electricity supply. Instead of tungsten, brass was used as leaf material to make the mechanical MLC more affordable. The physical properties and clinical applicability of this mechanical MLC are presented here. The leakage strongly depends on the fieldsize of the therapy unit due to scatter effects. The maximum transmission through the leaves measured 2.5 cm from the end-to-end gap, within a field size of 20 cm × 30 cm defined by jaws of the therapy unit at 80 cm SAD, amounts 4.2%, normalized to an open 10 cm × 10 cm field, created by the mechanical MLC. Within a precollimated field size of 12.5 cm × 12.5 cm, the end-to-end leakage is 6.5% normalized to an open 10 cm × 10 cm field as well. This characteristic is clinically acceptable considering the criteria for non-IMRT MLCs of the International Electrotechnical Commission (IEC 60601-2-1). The penumbra for a 10 cm × 10 cm field was measured to be 9.14 mm in plane and 8.38 mm cross plane. The clinical applicability of the designed mechanical MLC was affirmed by measurements relating to all relevant clinical properties such as penumbra, leakage, output factors and field widths. Hence this novel device presents an apt way forward to make radiotherapy with conformal fields possible in low-infrastructure environments, using gantry based Co-60 therapy units.
Poster — Thur Eve — 19: Performance assessment of a 160-leaf beam collimation system
DOE Office of Scientific and Technical Information (OSTI.GOV)
Ali, E. S. M.; La Russa, D. J.; Vandervoort, E.
2014-08-15
In this study, the performance of the new beam collimation system with 160 leaves, each with a 5 mm leaf width projected at isocenter, is evaluated in terms of positional accuracy and plan/delivery quality. Positional accuracy was evaluated using a set of static and dynamic MLC/jaw delivery patterns at different gantry angles, dose rates, and MLC/jaw speeds. The impact on IMRT plan quality was assessed by comparing against a previous generation collimation system using the same optimization parameters, while delivery quality was quantified using a combination of patient-specific QA measurements with ion chambers, film, and a bi-planar diode array. Positionalmore » accuracy for four separate units was comparable. The field size accuracy, junction width, and total displacement over 16 cm leaf travel are 0.3 ± 0.2 mm, 0.4 ± 0.3 mm, and 0.5 ± 0.2 mm, respectively. The typical leaf minor offset is 0.05 ± 0.04 mm, and MLC hysteresis effects are 0.2 ± 0.1 mm over 16 cm travel. The dynamic output is linear with MU and MLC/jaw speed, and is within 0.7 ± 0.3 % of the planning system value. Plan quality is significantly improved both in terms of target coverage and OAR sparing due, in part, to the larger allowable MLC and jaw speeds. γ-index pass rates for the patient-specific QA measurements exceeded 97% using criteria of 2%/2 mm. In conclusion, the performance of the Agility system is consistent among four separate installations, and is superior to its previous generations of collimation systems.« less
NASA Astrophysics Data System (ADS)
Sung, Changhyuck; Lim, Seokjae; Kim, Hyungjun; Kim, Taesu; Moon, Kibong; Song, Jeonghwan; Kim, Jae-Joon; Hwang, Hyunsang
2018-03-01
To improve the classification accuracy of an image data set (CIFAR-10) by using analog input voltage, synapse devices with excellent conductance linearity (CL) and multi-level cell (MLC) characteristics are required. We analyze the CL and MLC characteristics of TaOx-based filamentary resistive random access memory (RRAM) to implement the synapse device in neural network hardware. Our findings show that the number of oxygen vacancies in the filament constriction region of the RRAM directly controls the CL and MLC characteristics. By adopting a Ta electrode (instead of Ti) and the hot-forming step, we could form a dense conductive filament. As a result, a wide range of conductance levels with CL is achieved and significantly improved image classification accuracy is confirmed.
Actin-myosin-based contraction is responsible for apoptotic nuclear disintegration.
Croft, Daniel R; Coleman, Mathew L; Li, Shuixing; Robertson, David; Sullivan, Teresa; Stewart, Colin L; Olson, Michael F
2005-01-17
Membrane blebbing during the apoptotic execution phase results from caspase-mediated cleavage and activation of ROCK I. Here, we show that ROCK activity, myosin light chain (MLC) phosphorylation, MLC ATPase activity, and an intact actin cytoskeleton, but not microtubular cytoskeleton, are required for disruption of nuclear integrity during apoptosis. Inhibition of ROCK or MLC ATPase activity, which protect apoptotic nuclear integrity, does not affect caspase-mediated degradation of nuclear proteins such as lamins A, B1, or C. The conditional activation of ROCK I was sufficient to tear apart nuclei in lamin A/C null fibroblasts, but not in wild-type fibroblasts. Thus, apoptotic nuclear disintegration requires actin-myosin contractile force and lamin proteolysis, making apoptosis analogous to, but distinct from, mitosis where nuclear disintegration results from microtubule-based forces and from lamin phosphorylation and depolymerization.
Actin-myosin–based contraction is responsible for apoptotic nuclear disintegration
Croft, Daniel R.; Coleman, Mathew L.; Li, Shuixing; Robertson, David; Sullivan, Teresa; Stewart, Colin L.; Olson, Michael F.
2005-01-01
Membrane blebbing during the apoptotic execution phase results from caspase-mediated cleavage and activation of ROCK I. Here, we show that ROCK activity, myosin light chain (MLC) phosphorylation, MLC ATPase activity, and an intact actin cytoskeleton, but not microtubular cytoskeleton, are required for disruption of nuclear integrity during apoptosis. Inhibition of ROCK or MLC ATPase activity, which protect apoptotic nuclear integrity, does not affect caspase-mediated degradation of nuclear proteins such as lamins A, B1, or C. The conditional activation of ROCK I was sufficient to tear apart nuclei in lamin A/C null fibroblasts, but not in wild-type fibroblasts. Thus, apoptotic nuclear disintegration requires actin-myosin contractile force and lamin proteolysis, making apoptosis analogous to, but distinct from, mitosis where nuclear disintegration results from microtubule-based forces and from lamin phosphorylation and depolymerization. PMID:15657395
Paleomagnetic dating: Methods, MATLAB software, example
NASA Astrophysics Data System (ADS)
Hnatyshin, Danny; Kravchinsky, Vadim A.
2014-09-01
A MATLAB software tool has been developed to provide an easy to use graphical interface for the plotting and interpretation of paleomagnetic data. The tool takes either paleomagnetic directions or paleopoles and compares them to a user defined apparent polar wander path or secular variation curve to determine the age of a paleomagnetic sample. Ages can be determined in two ways, either by translating the data onto the reference curve, or by rotating it about a set location (e.g. sampling location). The results are then compiled in data tables which can be exported as an excel file. This data can also be plotted using variety of built-in stereographic projections, which can then be exported as an image file. This software was used to date the giant Sukhoi Log gold deposit in Russia. Sukhoi Log has undergone a complicated history of faulting, folding, metamorphism, and is the vicinity of many granitic bodies. Paleomagnetic analysis of Sukhoi Log allowed for the timing of large scale thermal or chemical events to be determined. Paleomagnetic analysis from gold mineralized black shales was used to define the natural remanent magnetization recorded at Sukhoi Log. The obtained paleomagnetic direction from thermal demagnetization produced a paleopole at 61.3°N, 155.9°E, with the semi-major axis and semi-minor axis of the 95% confidence ellipse being 16.6° and 15.9° respectively. This paleopole is compared to the Siberian apparent polar wander path (APWP) by translating the paleopole to the nearest location on the APWP. This produced an age of 255.2- 31.0+ 32.0Ma and is the youngest well defined age known for Sukhoi Log. We propose that this is the last major stage of activity at Sukhoi Log, and likely had a role in determining the present day state of mineralization seen at the deposit.
Interface Message Processors for the ARPA Computer Network
1975-04-01
Pluribus IMP construction and checkout; sizeable changes to the i*4P message-processing algorithms: and Satellite IMP issues. The IMP message...extremely low cost modification design. We have begun to consider changes to the MLC design which would enable the MLC to suppress continuous breaks...existing authentication mechanisms need not make these changes . 2.7 Other Topics During the first quarter BBN constructed an environmental test chamber
Kidney protection against ischemia/reperfusion injury by myofibrillogenesis regulator-1.
Wang, Xiaoreng; Tao, Tianqi; Ding, Rui; Song, Dandan; Liu, Mi; Xie, Yuansheng; Liu, Xiuhua
2014-01-01
Ischemia/reperfusion (I/R) injury is characterized by cytoskeletal reorganization and loss of polarity in proximal tubule epithelial cells. Previously, we showed that myofibrillogenesis regulator (MR)-1 promoted actin organization in cardiomyocytes. MR-1 is also expressed in the kidney. In this study, we investigated MR-1 expression in acute renal failure induced by I/R in Sprague-Dawley rats. We determined the MR-1 expression and the ratio of fibrous actin (F-actin) to globular actin (G-actin). HK-2 cells were treated with or without hypoxia/reoxygenation (H/R), and MR-1 levels were increased by adenoviral overexpression or silenced by RNA interference. I/R and H/R resulted in cellular injury and decreases of MR-1, the F-/G-actin ratio, and myosin light chain (MLC)-2. MR-1 overexpression attenuated H/R-induced cell injury and loss of surface membrane polarity of actin. MR-1 overexpression also increased the expression and phosphorylation of MLC-2 and MLC kinase, which were decreased in MR-1-silenced and H/R-treated cells. Together, these data show that MR-1 promoted actin polarity on the membrane surface and protected HK-2 cells from H/R injury. The mechanism might involve the rapid organization of F-actin through the upregulation and phosphorylation of MLC-2.
Development and validation of a BEAMnrc component module for a miniature multileaf collimator.
Doerner, E; Hartmann, G H
2012-05-21
A new component module (CM) named mini multileaf collimator (mMLC) was developed for the Monte Carlo code BEAMnrc. It models the geometry of the add-on miniature multileaf collimator ModuLeaf (MRC Systems GmbH, Heidelberg, Germany, now part of Siemens, Erlangen, Germany). The new CM is partly based on the existing CM called DYNVMLC. The development was performed using a modified EGSnrc platform which enables us to work in the Microsoft Visual Studio environment. In order to validate the new CM, the PRIMUS linac with 6 MV x-rays (Siemens OCS, Concord, CA, USA) equipped with the ModuLeaf mMLC was modelled. Validation was performed by two methods: (a) a ray-tracing method to check the correct geometry of the multileaf collimator (MLC) and (b) a comparison of calculated and measured results of the following dosimetrical parameters: output factors, dose profiles, field edge position penumbra, MLC interleaf leakage and transmission values. Excellent agreement was found for all parameters. It was, in particular, found that the relationship between leaf position and field edge depending on the shape of the leaf ends can be investigated with a higher accuracy by this new CM than by measurements demonstrating the usefulness of the new CM.
Development and validation of a BEAMnrc component module for a miniature multileaf collimator
NASA Astrophysics Data System (ADS)
Doerner, E.; Hartmann, G. H.
2012-05-01
A new component module (CM) named mini multileaf collimator (mMLC) was developed for the Monte Carlo code BEAMnrc. It models the geometry of the add-on miniature multileaf collimator ModuLeaf (MRC Systems GmbH, Heidelberg, Germany, now part of Siemens, Erlangen, Germany). The new CM is partly based on the existing CM called DYNVMLC. The development was performed using a modified EGSnrc platform which enables us to work in the Microsoft Visual Studio environment. In order to validate the new CM, the PRIMUS linac with 6 MV x-rays (Siemens OCS, Concord, CA, USA) equipped with the ModuLeaf mMLC was modelled. Validation was performed by two methods: (a) a ray-tracing method to check the correct geometry of the multileaf collimator (MLC) and (b) a comparison of calculated and measured results of the following dosimetrical parameters: output factors, dose profiles, field edge position penumbra, MLC interleaf leakage and transmission values. Excellent agreement was found for all parameters. It was, in particular, found that the relationship between leaf position and field edge depending on the shape of the leaf ends can be investigated with a higher accuracy by this new CM than by measurements demonstrating the usefulness of the new CM.
Aspects of cooling tower biocides and protozoa
DOE Office of Scientific and Technical Information (OSTI.GOV)
Berk, S.G.; Ashburn, R.J.; Ting, R.S.
1998-12-31
Previous work has shown that certain cooling tower amoebae and ciliated protozoa are resistant to several cooling tower biocides, even at the manufacturer`s recommended dosages. For the present study, an Acunthumoeba species was isolated from a cooling tower in Australia. Suspensions of the trophozoites (feeding stages) were exposed to isothiazolones. Cysts were tested separately. The minimum lethal concentration (MLC) for trophozoites was between 31-62 ppm of the biocide product, which is slightly less than the MLC for an amoebae species from the United States; and cyst forms were twofold more resistant than those of the US species, with a MLCmore » of 62,500 ppm. A ciliate and an amoeba species were also exposed to bromochlorodimethylhydantoin. The MLC for the ciliate species was 1 ppm of the biocide product, and the MLC was 30--40 ppm for the amoeba trophozoites. Since amoebae can expel vesicles containing live Legionella, experiments were conducted to determine whether exposure of Acunthamoebu polyphugu to biocides influenced release of such potentially infectious particles. Vesicle release was not inhibited by any of the three biocides: quaternary ammonium compounds (QACs), isothiazolones, and a thiocarbamate compound. These results suggest that amoebae from various sources are resistant to recommended levels of biocides, and the amoebae may continue to release potentially infectious vesicles in the presence of biocides.« less
75 FR 76426 - Privacy Act of 1974; System of Records
Federal Register 2010, 2011, 2012, 2013, 2014
2010-12-08
..., access control lists, file system permissions, intrusion detection and prevention systems and log..., address, mailing address, country, organization, phone, fax, mobile, pager, Defense Switched Network (DSN..., address, mailing address, country, organization, phone, fax, mobile, pager, Defense Switched Network (DSN...
An Efficient Format for Nearly Constant-Time Access to Arbitrary Time Intervals in Large Trace Files
Chan, Anthony; Gropp, William; Lusk, Ewing
2008-01-01
A powerful method to aid in understanding the performance of parallel applications uses log or trace files containing time-stamped events and states (pairs of events). These trace files can be very large, often hundreds or even thousands of megabytes. Because of the cost of accessing and displaying such files, other methods are often used that reduce the size of the tracefiles at the cost of sacrificing detail or other information. This paper describes a hierarchical trace file format that provides for display of an arbitrary time window in a time independent of the total size of the file and roughlymore » proportional to the number of events within the time window. This format eliminates the need to sacrifice data to achieve a smaller trace file size (since storage is inexpensive, it is necessary only to make efficient use of bandwidth to that storage). The format can be used to organize a trace file or to create a separate file of annotations that may be used with conventional trace files. We present an analysis of the time to access all of the events relevant to an interval of time and we describe experiments demonstrating the performance of this file format.« less
A Scientific Data Provenance Harvester for Distributed Applications
DOE Office of Scientific and Technical Information (OSTI.GOV)
Stephan, Eric G.; Raju, Bibi; Elsethagen, Todd O.
Data provenance provides a way for scientists to observe how experimental data originates, conveys process history, and explains influential factors such as experimental rationale and associated environmental factors from system metrics measured at runtime. The US Department of Energy Office of Science Integrated end-to-end Performance Prediction and Diagnosis for Extreme Scientific Workflows (IPPD) project has developed a provenance harvester that is capable of collecting observations from file based evidence typically produced by distributed applications. To achieve this, file based evidence is extracted and transformed into an intermediate data format inspired in part by W3C CSV on the Web recommendations, calledmore » the Harvester Provenance Application Interface (HAPI) syntax. This syntax provides a general means to pre-stage provenance into messages that are both human readable and capable of being written to a provenance store, Provenance Environment (ProvEn). HAPI is being applied to harvest provenance from climate ensemble runs for Accelerated Climate Modeling for Energy (ACME) project funded under the U.S. Department of Energy’s Office of Biological and Environmental Research (BER) Earth System Modeling (ESM) program. ACME informally provides provenance in a native form through configuration files, directory structures, and log files that contain success/failure indicators, code traces, and performance measurements. Because of its generic format, HAPI is also being applied to harvest tabular job management provenance from Belle II DIRAC scheduler relational database tables as well as other scientific applications that log provenance related information.« less
Federal Register 2010, 2011, 2012, 2013, 2014
2012-02-22
... visitors are required to show photographic identification, pass through a metal detector, and sign the EPA visitor log. All visitor bags are processed through an X-ray machine and subject to search. Visitors will...
Dynamic temperature response of electrocaloric multilayer capacitors
NASA Astrophysics Data System (ADS)
Kwon, Beomjin; Roh, Im-Jun; Baek, Seung-Hyub; Keun Kim, Seong; Kim, Jin-Sang; Kang, Chong-Yun
2014-05-01
We measure and model the dynamic temperature response of electrocaloric (EC) multilayer capacitors (MLCs) which have been recently highlighted as novel solid-state refrigerators. The MLC temperature responses depend on the operation voltage waveform, thus we consider three types of voltage waveforms, which include square, triangular, and trapezoidal. Further, to implement an effective refrigeration cycle, the waveform frequency and duty cycle should be carefully chosen. First, our model is fitted to the measurements to evaluate an effective EC power and thermal properties, and calculates an effective cooling power for an EC MLC. The prediction shows that for a MLC with a thermal relaxation time for cooling, trc, a square voltage waveform with a duty cycle of 0 < d ≤ 0.3 and a period of trc < P ≤ 1.4trc provides the maximum cooling power. This work will help to improve the implementing methods for EC refrigeration cycles.
Neuropsychological constraints to human data production on a global scale
NASA Astrophysics Data System (ADS)
Gros, C.; Kaczor, G.; Marković, D.
2012-01-01
Which are the factors underlying human information production on a global level? In order to gain an insight into this question we study a corpus of 252-633 mil. publicly available data files on the Internet corresponding to an overall storage volume of 284-675 Terabytes. Analyzing the file size distribution for several distinct data types we find indications that the neuropsychological capacity of the human brain to process and record information may constitute the dominant limiting factor for the overall growth of globally stored information, with real-world economic constraints having only a negligible influence. This supposition draws support from the observation that the files size distributions follow a power law for data without a time component, like images, and a log-normal distribution for multimedia files, for which time is a defining qualia.
SU-F-T-604: Dosimetric Evaluation of Intracranial Stereotactic Radiotherapy Plans On a LINAC
DOE Office of Scientific and Technical Information (OSTI.GOV)
Sheth, N; Tabibian, A; Rose, J
2016-06-15
Purpose: To evaluate the dosimetry of cranial stereotactic radiotherapy (SRT) plans of varying techniques on linac that meets appropriate TG-142 tolerances using 1 cm leaf width multileaf collimator (MLC). Methods: Seventeen spherical targets were generated in the center of a head phantom with diameters ranging 8 mm to 40 mm. SRT plans used 100° non-coplanar arcs and 5 couch angles with 35° spacing. The field size was target plus 1 mm margin. Four plans were created for each target: symmetrical jaws blocking for 5 arcs with 0° collimator (J1C), symmetrical jaws blocking with 5 clockwise arcs with 0° collimator andmore » 5 counter-clockwise arcs with 45° collimator (J2C), MLC blocking for 5 dynamic conformal arcs with 0° collimator (M1C), and MLC blocking for 5 clockwise dynamic conformal arcs with 0° collimators and 5 counter-clockwise dynamic conformal arcs with 45° collimator (M2C).Conformity was evaluated using a ratio of Rx to target volume (PITV). Heterogeneity was determined using a ratio of maximum dose to Rx dose. Falloff was scored using CGIg: difference of effective radii of spheres equal to half and full Rx volumes. Results: All plans met RTOG SRS criteria for conformity and heterogeneity. The mean PITV was 1.52±0.07, 1.49±0.08, 1.39±0.05, and 1.37±0.04 for J1C, J2C, M1C, and M2C plans respectively. The mean CGIg was 75.35±15.79, 74.19±16.66, 77.14±15.12, and 76.28±15.78 for J1C, J2C, M1C, and M2C plans respectively. The mean MDPD was 1.25±0.00 for all techniques. Conclusion: Clinically acceptable SRT plans for spherical targets were created on a linac with 1 cm MLC. Adding two collimator angles and MLC to arcs each improved conformity. The MLC improved the dose falloff while two collimator angles degraded it. This technique can expand the availability of SRT to patients especially to those who cannot travel to a facility with a dedicated stereotactic radiosurgery machine.« less
Magnetic decoupling of the linac in a low field biplanar linac-MR system.
St Aubin, J; Steciw, S; Fallone, B G
2010-09-01
The integration of a low field biplanar magnetic resonance (MR) imager and linear accelerator (linac) causes magnetic interference at the linac due to the MR fringe fields. In order to eliminate this interference, passive and active magnetic shielding designs are investigated. The optimized design of passive magnetic shielding was performed using the finite element method. The design was required to achieve no greater than a 20% electron beam loss within the linac waveguide and electron gun, no greater than 0.06 T at the multileaf collimator (MLC) motors, and generate a distortion of the main MR imaging volume of no greater than 300 ppm. Through the superposition of the analytical solution for a single current carrying wire loop, active shielding designs in the form of three and four sets of coil pairs surrounding the linac waveguide and electron gun were also investigated. The optimized current and coil center locations that yielded the best cancellation of the MR fringe fields at the linac were determined using sequential quadratic programming. Optimized passive shielding in the form of two steel cylinders was designed to meet the required constraints. When shielding the MLC motors along with the waveguide and electron gun, the thickness of the cylinders was less than 1 mm. If magnetically insensitive MLC motors are used, no MLC shielding would be required and the waveguide shield (shielding the waveguide and electron gun) became 1.58 mm thick. In addition, the optimized current and coil spacing for active shielding was determined for both three and four coil pair configurations. The results of the active shielding optimization produced no beam loss within the waveguide and electron gun and a maximum MR field distortion of 91 ppm over a 30 cm diameter spherical volume. Very simple passive and active shielding designs have been shown to magnetically decouple the linac from the MR imager in a low field biplanar linac-MR system. The MLC passive shielding produced the largest distortion of the MR field over the imaging volume. With the use of magnetically insensitive motors, the MR field distortion drops substantially since no MLC shield is required. The active shielding designs yielded no electron beam loss within the linac.
Sarkar, B; Pradhan, A; Munshi, A
2016-01-01
Linear accelerator (Linac) based stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT) using volumetric modulated arc therapy (VMAT) has been used for treating small intracranial lesions. Recent development in the Linacs such as inbuilt micro multileaf collimator (MLC) and flattening filter free (FFF) beam are intended to provide a better dose conformity and faster delivery when using VMAT technique. This study was aimed to compare the dosimetric outcomes and monitor units (MUs) of the stereotactic treatment plans for different commercially available MLC models and beam profiles. Ten patients having 12 planning target volume (PTV)/gross target volume's (GTVs) who received the SRS/SRT treatment in our clinic using Axesse Linac (considered reference arm gold standard) were considered for this study. The test arms comprised of plans using Elekta Agility with FFF, Elekta Agility with the plane beam, Elekta APEX, Varian Millennium 120, Varian Millennium 120HD, and Elekta Synergy in Monaco treatment planning system. Planning constraints and calculation grid spacing were not altered in the test plans. To objectively evaluate the efficacy of MLC-beam model, the resultant dosimetric outcomes were subtracted from the reference arm parameters. V95%, V100%, V105%, D1%, maximum dose, and mean dose of PTV/GTV showed a maximum inter MLC - beam model variation of 1.5% and 2% for PTV and GTV, respectively. Average PTV conformity index and heterogeneity index shows a variation in the range 0.56-0.63 and 1.08-1.11, respectively. Mean dose difference (excluding Axesse) for all organs varied between 1.1 cGy and 74.8 cGy (mean dose = 6.1 cGy standard deviation [SD] = 26.9 cGy) and 1.7 cGy-194.5 cGy (mean dose 16.1 cGy SD = 57.2 cGy) for single and multiple fraction, respectively. The dosimetry of VMAT-based SRS/SRT treatment plan had minimal dependence on MLC and beam model variations. All tested MLC and beam model could fulfil the desired PTV coverage and organs at risk dose constraints. The only notable difference was the halving of the MU for FFF beam as compared to the plane beam. This has the potential to reduce the total patient on couch time by 15% (approximately 2 min).
Sharma, Dayananda S; Jalali, Rakesh; Tambe, Chandrashekhar M; Animesh; Deshpande, Deepak D
2004-01-01
The aim of this work was to measure the dose to foetus both in vivo and in vitro during three-dimensional conformal radiation therapy (3DCRT) in a pregnant patient with a pituitary adenoma. The study was then extended to assess the components contributing to the foetal dose such as collimator scatter, internal scatter, head leakage, wedge scatter and multileaf collimator (MLC) effect. A 30-year-old pregnant woman with a non-functioning pituitary macroadenoma was planned for 3DCRT with 6MV X-ray using four equally weighted MLC-shaped non-coplanar wedged portals. In vivo dosimetry was carried out using thermoluminescent (TL) phosphor powder, which was placed at different positions on the patient, corresponding to different locations in the uterus and also at external os. In vitro measurements were also performed on a simulated phantom using the same set-up parameters and beam arrangement to verify the in vivo measured dose. Experiments were carried out to measure the respective contributions of different components towards peripheral dose. In vitro measured dose to foetus was found to be slightly more than that of in vivo measurement with a maximum of 0.044% of the prescribed dose of 45Gy, which corresponded to 0.0199+/-0.0008Gy. Thermoluminescence dosimeter (TLD) kept at the external os of the patient showed a dose of 0.031% of the prescribed dose. Among the various components of the peripheral dose (foetal dose) measured, head leakage was found to be the leading cause contributing 52%, followed by wedge scatter (31%), collimator scatter (14%) and internal scatter (13%). The use of MLC reduced not only the volume of normal brain irradiation as compared to open fields but also the peripheral dose by 10%. Radiotherapy of brain tumours during pregnancy poses a unique clinical situation and decisions to deliver radiotherapy should be taken after detailed in vitro and in vivo dosimetric measurements. Our findings suggest that the beam arrangement using 3-4-fields generally used for 3DCRT of brain tumour with MLC for optimal coverage can be employed for pregnant patients even in early trimester. A possible increase in foetal dose from wedges to a large extent can be compensated with the use of MLC.
Yoon, Jai-Woong; Sawant, Amit; Suh, Yelin; Cho, Byung-Chul; Suh, Tae-Suk; Keall, Paul
2011-07-01
In dynamic multileaf collimator (MLC) motion tracking with complex intensity-modulated radiation therapy (IMRT) fields, target motion perpendicular to the MLC leaf travel direction can cause beam holds, which increase beam delivery time by up to a factor of 4. As a means to balance delivery efficiency and accuracy, a moving average algorithm was incorporated into a dynamic MLC motion tracking system (i.e., moving average tracking) to account for target motion perpendicular to the MLC leaf travel direction. The experimental investigation of the moving average algorithm compared with real-time tracking and no compensation beam delivery is described. The properties of the moving average algorithm were measured and compared with those of real-time tracking (dynamic MLC motion tracking accounting for both target motion parallel and perpendicular to the leaf travel direction) and no compensation beam delivery. The algorithm was investigated using a synthetic motion trace with a baseline drift and four patient-measured 3D tumor motion traces representing regular and irregular motions with varying baseline drifts. Each motion trace was reproduced by a moving platform. The delivery efficiency, geometric accuracy, and dosimetric accuracy were evaluated for conformal, step-and-shoot IMRT, and dynamic sliding window IMRT treatment plans using the synthetic and patient motion traces. The dosimetric accuracy was quantified via a tgamma-test with a 3%/3 mm criterion. The delivery efficiency ranged from 89 to 100% for moving average tracking, 26%-100% for real-time tracking, and 100% (by definition) for no compensation. The root-mean-square geometric error ranged from 3.2 to 4.0 mm for moving average tracking, 0.7-1.1 mm for real-time tracking, and 3.7-7.2 mm for no compensation. The percentage of dosimetric points failing the gamma-test ranged from 4 to 30% for moving average tracking, 0%-23% for real-time tracking, and 10%-47% for no compensation. The delivery efficiency of moving average tracking was up to four times higher than that of real-time tracking and approached the efficiency of no compensation for all cases. The geometric accuracy and dosimetric accuracy of the moving average algorithm was between real-time tracking and no compensation, approximately half the percentage of dosimetric points failing the gamma-test compared with no compensation.
Venketasubramanian, Narayanaswamy; Lee, Chun Fan; Young, Sherry H; Tay, San San; Umapathi, Thirugnanam; Lao, Annabelle Y; Gan, Herminigildo H; Baroque Ii, Alejandro C; Navarro, Jose C; Chang, Hui Meng; Advincula, Joel M; Muengtaweepongsa, Sombat; Chan, Bernard P L; Chua, Carlos L; Wijekoon, Nirmala; de Silva, H Asita; Hiyadan, John Harold B; Suwanwela, Nijasri C; Wong, K S Lawrence; Poungvarin, Niphon; Eow, Gaik Bee; Chen, Christopher L H
2017-01-01
The Chinese Medicine NeuroAiD Efficacy on Stroke recovery - Extension (CHIMES-E) study is among the few acute stroke trials with long-term outcome data. We aimed to evaluate the recovery pattern and the influence of prognostic factors on treatment effect of MLC601 over 2 years. The CHIMES-E study evaluated the 2 years outcome of subjects aged ≥18 years with acute ischemic stroke, National Institutes of Health Stroke Scale (NIHSS) score 6-14, pre-stroke modified Rankin Scale (mRS) score ≤1 included in a multicenter, randomized, double-blind, placebo-controlled trial of MLC601 for 3 months. Standard stroke care and rehabilitation were allowed during follow-up with mRS score being assessed in-person at month (M) 3 and by telephone at M1, M6, M12, M18 and M24. Data from 880 subjects were analyzed. There was no difference in baseline characteristics between treatment groups. The proportion of subjects with mRS score 0-1 increased over time in favor of MLC601 most notably from M3 to M6, thereafter remaining stable up to M24, while the proportion deteriorating to mRS score ≥2 remained low at all time points. Older age (p < 0.01), female sex (p = 0.06), higher baseline NIHSS score (p < 0.01) and longer onset to treatment time (OTT; p < 0.01) were found to be predictors of poorer outcome at M3. Greater treatment effect, with more subjects improving on MLC601 than placebo, was seen among subjects with 2 or more prognostic factors (OR 1.65 at M3, 1.78 at M6, 1.90 at M12, 1.65 at M18, 1.39 at M24), especially in subjects with more severe stroke or longer OTT. The sustained benefits of MLC601 over 2 years were due to more subjects improving to functional independence at M6 and beyond compared to placebo. Selection of subjects with poorer prognosis, particularly those with more severe NIHSS score and longer OTT delay, as well as a long follow-up period, may improve the power of future trials investigating the treatment effect of neuroprotective or neurorestorative therapies. © 2016 The Author(s) Published by S. Karger AG, Basel.
Fast leaf-fitting with generalized underdose/overdose constraints for real-time MLC tracking
DOE Office of Scientific and Technical Information (OSTI.GOV)
Moore, Douglas, E-mail: douglas.moore@utsouthwestern.edu; Sawant, Amit; Ruan, Dan
2016-01-15
Purpose: Real-time multileaf collimator (MLC) tracking is a promising approach to the management of intrafractional tumor motion during thoracic and abdominal radiotherapy. MLC tracking is typically performed in two steps: transforming a planned MLC aperture in response to patient motion and refitting the leaves to the newly generated aperture. One of the challenges of this approach is the inability to faithfully reproduce the desired motion-adapted aperture. This work presents an optimization-based framework with which to solve this leaf-fitting problem in real-time. Methods: This optimization framework is designed to facilitate the determination of leaf positions in real-time while accounting for themore » trade-off between coverage of the PTV and avoidance of organs at risk (OARs). Derived within this framework, an algorithm is presented that can account for general linear transformations of the planned MLC aperture, particularly 3D translations and in-plane rotations. This algorithm, together with algorithms presented in Sawant et al. [“Management of three-dimensional intrafraction motion through real-time DMLC tracking,” Med. Phys. 35, 2050–2061 (2008)] and Ruan and Keall [Presented at the 2011 IEEE Power Engineering and Automation Conference (PEAM) (2011) (unpublished)], was applied to apertures derived from eight lung intensity modulated radiotherapy plans subjected to six-degree-of-freedom motion traces acquired from lung cancer patients using the kilovoltage intrafraction monitoring system developed at the University of Sydney. A quality-of-fit metric was defined, and each algorithm was evaluated in terms of quality-of-fit and computation time. Results: This algorithm is shown to perform leaf-fittings of apertures, each with 80 leaf pairs, in 0.226 ms on average as compared to 0.082 and 64.2 ms for the algorithms of Sawant et al., Ruan, and Keall, respectively. The algorithm shows approximately 12% improvement in quality-of-fit over the Sawant et al. approach, while performing comparably to Ruan and Keall. Conclusions: This work improves upon the quality of the Sawant et al. approach, but does so without sacrificing run-time performance. In addition, using this framework allows for complex leaf-fitting strategies that can be used to account for PTV/OAR trade-off during real-time MLC tracking.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Chengqiang, L; Yin, Y; Chen, L
Purpose: To investigate the impact of MLC position errors on simultaneous integrated boost intensity-modulated radiotherapy (SIB-IMRT) for patients with nasopharyngeal carcinoma. Methods: To compare the dosimetric differences between the simulated plans and the clinical plans, ten patients with locally advanced NPC treated with SIB-IMRT were enrolled in this study. All plans were calculated with an inverse planning system (Pinnacle3, Philips Medical System{sub )}. Random errors −2mm to 2mm{sub )},shift errors{sub (} 2mm,1mm and 0.5mm) and systematic extension/ contraction errors (±2mm, ±1mm and ±0.5mm) of the MLC leaf position were introduced respectively into the original plans to create the simulated plans.more » Dosimetry factors were compared between the original and the simulated plans. Results: The dosimetric impact of the random and system shift errors of MLC position was insignificant within 2mm, the maximum changes in D95% of PGTV,PTV1,PTV2 were-0.92±0.51%,1.00±0.24% and 0.62±0.17%, the maximum changes in the D0.1cc of spinal cord and brainstem were 1.90±2.80% and −1.78±1.42%, the maximum changes in the Dmean of parotids were1.36±1.23% and −2.25±2.04%.However,the impact of MLC extension or contraction errors was found significant. For 2mm leaf extension errors, the average changes in D95% of PGTV,PTV1,PTV2 were 4.31±0.67%,4.29±0.65% and 4.79±0.82%, the averaged value of the D0.1cc to spinal cord and brainstem were increased by 7.39±5.25% and 6.32±2.28%,the averaged value of the mean dose to left and right parotid were increased by 12.75±2.02%,13.39±2.17% respectively. Conclusion: The dosimetric effect was insignificant for random MLC leaf position errors up to 2mm. There was a high sensitivity to dose distribution for MLC extension or contraction errors.We should pay attention to the anatomic changes in target organs and anatomical structures during the course,individual radiotherapy was recommended to ensure adaptive doses.« less
Kaddour-Djebbar, I; Ansari, H R; Akhtar, R A; Abdel-Latif, A A
2005-01-01
There is evidence from our own laboratory and that of others that EP-receptor ligands are strong contractile agonists in bovine iris sphincter and that FP-receptor agonists are strong contractile agonists in cat iris sphincter. Here, we have investigated the effects of prostaglandin (PG) receptor agonists of the FP-, EP-, TP- and DP-class on myosin light chain (MLC) phosphorylation, p42/p44 MAP kinase phosphorylation and contraction in the iris sphincter of bovine and cat. Using three signal transduction mechanism assays, namely MLC phosphorylation, MAP kinase phosphorylation and contraction, we demonstrated that in bovine iris sphincter the rank order of potency of the PG agonists in the contractile and MLC phosphorylation assays is as follows: E2>U46619>F2alpha>D2, and in cat F2alpha>D2>E2>U46619. In the MAP kinase assay, in bovine iris sphincter the rank order of potency is E2>F2alpha and in cat F2alpha>E2. These conclusions are supported by the following findings: (1) In the contractile assay, in the bovine sphincter the EC50s for PGF2alpha, PGE2, U46619 and PGD2 were found to be 1.4x10(-7), 5.0x10(-9), 9.0x10(-9) and 1.3x10(-6)M, respectively, and the corresponding values in the cat were 1.9x10(-8), 2.3x10(-7), 1.5x10(-6) and 6.9x10(-8)M, respectively. (2) In the MLC phophorylation assay, in the bovine sphincter PGF2alpha, PGE2, U46619 and PGD2 increased MLC phophorylation by 118%, 165%, 153% and 72%, respectively, and the corresponding values in cat were 175%, 99%, 90% and 95%, respectively. (3) In the MAP kinase assay, in the bovine iris sphincter PGF2alpha and PGE2, increased MAP kinase phosphorylation by 276% and 328%, respectively, and the corresponding values in cat were 308% and 245%, respectively. The data presented demonstrate pronounced species differences in the effects of the prostanoids on the MLC kinase signaling pathway in bovine and cat irides and furthermore confirm the existence of FP-receptors in that of the bovine.
SU-E-T-100: Designing a QA Tool for Enhance Dynamic Wedges Based On Dynalog Files
DOE Office of Scientific and Technical Information (OSTI.GOV)
Yousuf, A; Hussain, A
2014-06-01
Purpose: A robust quality assurance (QA) program for computer controlled enhanced dynamic wedge (EDW) has been designed and tested. Calculations to perform such QA test is based upon the EDW dynamic log files generated during dose delivery. Methods: Varian record and verify system generates dynamic log (dynalog) files during dynamic dose delivery. The system generated dynalog files contain information such as date and time of treatment, energy, monitor units, wedge orientation, and type of treatment. It also contains the expected calculated segmented treatment tables (STT) and the actual delivered STT for the treatment delivery as a verification record. These filesmore » can be used to assess the integrity and precision of the treatment plan delivery. The plans were delivered with a 6 MV beam from a Varian linear accelerator. For available EDW angles (10°, 15°, 20°, 25°, 30°, 45°, and 60°) Varian STT values were used to manually calculate monitor units for each segment. It can also be used to calculate the EDW factors. Independent verification of fractional MUs per segment was performed against those generated from dynalog files. The EDW factors used to calculate MUs in TPS were dosimetrically verified in solid water phantom with semiflex chamber on central axis. Results: EDW factors were generated from the STT provided by Varian and verified against practical measurements. The measurements were in agreement of the order of 1 % to the calculated EDW data. Variation between the MUs per segment obtained from dynalog files and those manually calculated was found to be less than 2%. Conclusion: An efficient and easy tool to perform routine QA procedure of EDW is suggested. The method can be easily implemented in any institution without a need for expensive QA equipment. An error of the order of ≥2% can be easily detected.« less
75 FR 27051 - Privacy Act of 1974: System of Records
Federal Register 2010, 2011, 2012, 2013, 2014
2010-05-13
... address and appears below: DOT/FMCSA 004 SYSTEM NAME: National Consumer Complaint Database (NCCDB.... A system, database, and procedures for filing and logging consumer complaints relating to household... are stored in an automated system operated and maintained at the Volpe National Transportation Systems...