Sample records for patient setup verification

  1. Commissioning and quality assurance of an integrated system for patient positioning and setup verification in particle therapy.

    PubMed

    Pella, A; Riboldi, M; Tagaste, B; Bianculli, D; Desplanques, M; Fontana, G; Cerveri, P; Seregni, M; Fattori, G; Orecchia, R; Baroni, G

    2014-08-01

    In an increasing number of clinical indications, radiotherapy with accelerated particles shows relevant advantages when compared with high energy X-ray irradiation. However, due to the finite range of ions, particle therapy can be severely compromised by setup errors and geometric uncertainties. The purpose of this work is to describe the commissioning and the design of the quality assurance procedures for patient positioning and setup verification systems at the Italian National Center for Oncological Hadrontherapy (CNAO). The accuracy of systems installed in CNAO and devoted to patient positioning and setup verification have been assessed using a laser tracking device. The accuracy in calibration and image based setup verification relying on in room X-ray imaging system was also quantified. Quality assurance tests to check the integration among all patient setup systems were designed, and records of daily QA tests since the start of clinical operation (2011) are presented. The overall accuracy of the patient positioning system and the patient verification system motion was proved to be below 0.5 mm under all the examined conditions, with median values below the 0.3 mm threshold. Image based registration in phantom studies exhibited sub-millimetric accuracy in setup verification at both cranial and extra-cranial sites. The calibration residuals of the OTS were found consistent with the expectations, with peak values below 0.3 mm. Quality assurance tests, daily performed before clinical operation, confirm adequate integration and sub-millimetric setup accuracy. Robotic patient positioning was successfully integrated with optical tracking and stereoscopic X-ray verification for patient setup in particle therapy. Sub-millimetric setup accuracy was achieved and consistently verified in daily clinical operation.

  2. Marker-based quantification of interfractional tumor position variation and the use of markers for setup verification in radiation therapy for esophageal cancer.

    PubMed

    Jin, Peng; van der Horst, Astrid; de Jong, Rianne; van Hooft, Jeanin E; Kamphuis, Martijn; van Wieringen, Niek; Machiels, Melanie; Bel, Arjan; Hulshof, Maarten C C M; Alderliesten, Tanja

    2015-12-01

    The aim of this study was to quantify interfractional esophageal tumor position variation using markers and investigate the use of markers for setup verification. Sixty-five markers placed in the tumor volumes of 24 esophageal cancer patients were identified in computed tomography (CT) and follow-up cone-beam CT. For each patient we calculated pairwise distances between markers over time to evaluate geometric tumor volume variation. We then quantified marker displacements relative to bony anatomy and estimated the variation of systematic (Σ) and random errors (σ). During bony anatomy-based setup verification, we visually inspected whether the markers were inside the planning target volume (PTV) and attempted marker-based registration. Minor time trends with substantial fluctuations in pairwise distances implied tissue deformation. Overall, Σ(σ) in the left-right/cranial-caudal/anterior-posterior direction was 2.9(2.4)/4.1(2.4)/2.2(1.8) mm; for the proximal stomach, it was 5.4(4.3)/4.9(3.2)/1.9(2.4) mm. After bony anatomy-based setup correction, all markers were inside the PTV. However, due to large tissue deformation, marker-based registration was not feasible. Generally, the interfractional position variation of esophageal tumors is more pronounced in the cranial-caudal direction and in the proximal stomach. Currently, marker-based setup verification is not feasible for clinical routine use, but markers can facilitate the setup verification by inspecting whether the PTV covers the tumor volume adequately. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  3. SU-E-J-22: A Feasibility Study On KV-Based Whole Breast Radiation Patient Setup

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Huang, Q; Zhang, M; Yue, N

    Purpose: In room kilovoltage x-ray (kV) imaging provides higher contrast than Megavoltage (MV) imaging with faster acquisition time compared with on-board cone-beam computed tomography (CBCT), thus improving patient setup accuracy and efficiency. In this study we evaluated the clinical feasibility of utilizing kV imaging for whole breast radiation patient setup. Methods: For six breast cancer patients with whole breast treatment plans using two opposed tangential fields, MV-based patient setup was conducted by aligning patient markers with in room lasers and MV portal images. Beam-eye viewed kV images were acquired using Varian OBI system after the set up process. In housemore » software was developed to transfer MLC blocks information overlaying onto kV images to demonstrate the field shape for verification. KV-based patient digital shift was derived by performing rigid registration between kV image and the digitally reconstructed radiography (DRR) to align the bony structure. This digital shift between kV-based and MV-based setup was defined as setup deviation. Results: Six sets of kV images were acquired for breast patients. The mean setup deviation was 2.3mm, 2.2mm and 1.8mm for anterior-posterior, superior-inferior and left-right direction respectively. The average setup deviation magnitude was 4.3±1.7mm for six patients. Patient with large breast had a larger setup deviation (4.4–6.2mm). There was no strong correlation between MV-based shift and setup deviation. Conclusion: A preliminary clinical workflow for kV-based whole breast radiation setup was established and tested. We observed setup deviation of the magnitude below than 5mm. With the benefit of providing higher contrast and MLC block overlaid on the images for treatment field verification, it is feasible to use kV imaging for breast patient setup.« less

  4. SU-E-J-34: Setup Accuracy in Spine SBRT Using CBCT 6D Image Guidance in Comparison with 6D ExacTrac

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Han, Z; Yip, S; Lewis, J

    2015-06-15

    Purpose Volumetric information of the spine captured on CBCT can potentially improve the accuracy in spine SBRT setup that has been commonly performed through 2D radiographs. This work evaluates the setup accuracy in spine SBRT using 6D CBCT image guidance that recently became available on Varian systems. Methods ExacTrac radiographs have been commonly used for Spine SBRT setup. The setup process involves first positioning patients with lasers followed by localization imaging, registration, and repositioning. Verification images are then taken providing the residual errors (ExacTracRE) before beam on. CBCT verification is also acquired in our institute. The availability of both ExacTracmore » and CBCT verifications allows a comparison study. 41 verification CBCT of 16 patients were retrospectively registered with the planning CT enabling 6D corrections, giving CBCT residual errors (CBCTRE) which were compared with ExacTracRE. Results The RMS discrepancies between CBCTRE and ExacTracRE are 1.70mm, 1.66mm, 1.56mm in vertical, longitudinal and lateral directions and 0.27°, 0.49°, 0.35° in yaw, roll and pitch respectively. The corresponding mean discrepancies (and standard deviation) are 0.62mm (1.60mm), 0.00mm (1.68mm), −0.80mm (1.36mm) and 0.05° (0.58°), 0.11° (0.48°), −0.16° (0.32°). Of the 41 CBCT, 17 had high-Z surgical implants. No significant difference in ExacTrac-to-CBCT discrepancy was observed between patients with and without the implants. Conclusion Multiple factors can contribute to the discrepancies between CBCT and ExacTrac: 1) the imaging iso-centers of the two systems, while calibrated to coincide, can be different; 2) the ROI used for registration can be different especially if ribs were included in ExacTrac images; 3) small patient motion can occur between the two verification image acquisitions; 4) the algorithms can be different between CBCT (volumetric) and ExacTrac (radiographic) registrations.« less

  5. SU-E-J-245: Is Off-Line Adaptive Radiotherapy Sufficient for Head and Neck Cancer with IGRT?

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Li, Z; Cleveland Clinic, Cleveland, OH; Shang, Q

    2014-06-01

    Purpose: Radiation doses delivered to patients with head and neck cancer (HN) may deviate from the planned doses because of variations in patient setup and anatomy. This study was to evaluate whether off-line Adaptive Radiotherapy (ART) is sufficient. Methods: Ten HN patients, who received IMRT under daily imaging guidance using CT-on-rail/KV-CBCT, were randomly selected for this study. For each patient, the daily treatment setup was corrected with translational only directions. Sixty weekly verification CTs were retrospectively analyzed. On these weekly verification CTs, the tumor volumes and OAR contours were manually delineated by a physician. With the treatment iso-center placed onmore » the verification CTs, according to the recorded clinical shifts, the treatment beams from the original IMRT plans were then applied to these CTs to calculate the delivered doses. The electron density of the planning CTs and weekly CTs were overridden to 1 g/cm3. Results: Among 60 fractions, D99 of the CTVs in 4 fractions decreased more than 5% of the planned doses. The maximum dose of the spinal cord exceeded 10% of the planned values in 2 fractions. A close examination indicated that the dose discrepancy in these 6 fractions was due to patient rotations, especially shoulder rotations. After registering these 6 CTs with the planning CT allowing six degree of freedoms, the maximum rotations around 3 axes were > 1.5° for these fractions. With rotation setup errors removed, 4 out of 10 patients still required off-line ART to accommodate anatomical changes. Conclusion: A significant shoulder rotations were observed in 10% fractions, requiring patient re-setup. Off-line ART alone is not sufficient to correct for random variations of patient position, although ART is effective to adapt to patients' gradual anatomic changes. Re-setup or on-line ART may be considered for patients with large deviations detected early by daily IGRT images. The study is supported in part by Siemens Medical Solutions.« less

  6. An Optimized Online Verification Imaging Procedure for External Beam Partial Breast Irradiation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Willis, David J., E-mail: David.Willis@petermac.or; Royal Melbourne Institute of Technology University, Melbourne, Victoria; Kron, Tomas

    2011-07-01

    The purpose of this study was to evaluate the capabilities of a kilovoltage (kV) on-board imager (OBI)-equipped linear accelerator in the setting of on-line verification imaging for external-beam partial breast irradiation. Available imaging techniques were optimized and assessed for image quality using a modified anthropomorphic phantom. Imaging dose was also assessed. Imaging techniques were assessed for physical clearance between patient and treatment machine using a volunteer. Nonorthogonal kV image pairs were identified as optimal in terms of image quality, clearance, and dose. After institutional review board approval, this approach was used for 17 patients receiving accelerated partial breast irradiation. Imagingmore » was performed before every fraction verification with online correction of setup deviations >5 mm (total image sessions = 170). Treatment staff rated risk of collision and visibility of tumor bed surgical clips where present. Image session duration and detected setup deviations were recorded. For all cases, both image projections (n = 34) had low collision risk. Surgical clips were rated as well as visualized in all cases where they were present (n = 5). The average imaging session time was 6 min, 16 sec, and a reduction in duration was observed as staff became familiar with the technique. Setup deviations of up to 1.3 cm were detected before treatment and subsequently confirmed offline. Nonorthogonal kV image pairs allowed effective and efficient online verification for partial breast irradiation. It has yet to be tested in a multicenter study to determine whether it is dependent on skilled treatment staff.« less

  7. MO-F-CAMPUS-T-03: Data Driven Approaches for Determination of Treatment Table Tolerance Values for Record and Verification Systems

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Gupta, N; DiCostanzo, D; Fullenkamp, M

    2015-06-15

    Purpose: To determine appropriate couch tolerance values for modern radiotherapy linac R&V systems with indexed patient setup. Methods: Treatment table tolerance values have been the most difficult to lower, due to many factors including variations in patient positioning and differences in table tops between machines. We recently installed nine linacs with similar tables and started indexing every patient in our clinic. In this study we queried our R&V database and analyzed the deviation of couch position values from the acquired values at verification simulation for all patients treated with indexed positioning. Mean and standard deviations of daily setup deviations weremore » computed in the longitudinal, lateral and vertical direction for 343 patient plans. The mean, median and standard error of the standard deviations across the whole patient population and for some disease sites were computed to determine tolerance values. Results: The plot of our couch deviation values showed a gaussian distribution, with some small deviations, corresponding to setup uncertainties on non-imaging days, and SRS/SRT/SBRT patients, as well as some large deviations which were spot checked and found to be corresponding to indexing errors that were overriden. Setting our tolerance values based on the median + 1 standard error resulted in tolerance values of 1cm lateral and longitudinal, and 0.5 cm vertical for all non- SRS/SRT/SBRT cases. Re-analizing the data, we found that about 92% of the treated fractions would be within these tolerance values (ignoring the mis-indexed patients). We also analyzed data for disease site based subpopulations and found no difference in the tolerance values that needed to be used. Conclusion: With the use of automation, auto-setup and other workflow efficiency tools being introduced into radiotherapy workflow, it is very essential to set table tolerances that allow safe treatments, but flag setup errors that need to be reassessed before treatments.« less

  8. Accurate setup of paraspinal patients using a noninvasive patient immobilization cradle and portal imaging.

    PubMed

    Lovelock, D Michael; Hua, Chiaho; Wang, Ping; Hunt, Margie; Fournier-Bidoz, Nathalie; Yenice, Kamil; Toner, Sean; Lutz, Wendell; Amols, Howard; Bilsky, Mark; Fuks, Zvi; Yamada, Yoshiya

    2005-08-01

    Because of the proximity of the spinal cord, effective radiotherapy of paraspinal tumors to high doses requires highly conformal dose distributions, accurate patient setup, setup verification, and patient immobilization. An immobilization cradle has been designed to facilitate the rapid setup and radiation treatment of patients with paraspinal disease. For all treatments, patients were set up to within 2.5 mm of the design using an amorphous silicon portal imager. Setup reproducibility of the target using the cradle and associated clinical procedures was assessed by measuring the setup error prior to any correction. From 350 anterior/posterior images, and 303 lateral images, the standard deviations, as determined by the imaging procedure, were 1.3 m, 1.6 m, and 2.1 in the ant/post, right/left, and superior/inferior directions. Immobilization was assessed by measuring patient shifts between localization images taken before and after treatment. From 67 ant/post image pairs and 49 lateral image pairs, the standard deviations were found to be less than 1 mm in all directions. Careful patient positioning and immobilization has enabled us to develop a successful clinical program of high dose, conformal radiotherapy of paraspinal disease using a conventional Linac equipped with dynamic multileaf collimation and an amorphous silicon portal imager.

  9. A new verification film system for routine quality control of radiation fields: Kodak EC-L.

    PubMed

    Hermann, A; Bratengeier, K; Priske, A; Flentje, M

    2000-06-01

    The use of modern irradiation techniques requires better verification films for determining set-up deviations and patient movements during the course of radiation treatment. This is an investigation of the image quality and time requirement of a new verification film system compared to a conventional portal film system. For conventional verifications we used Agfa Curix HT 1000 films which were compared to the new Kodak EC-L film system. 344 Agfa Curix HT 1000 and 381 Kodak EC-L portal films of different tumor sites (prostate, rectum, head and neck) were visually judged on a light box by 2 experienced physicians. Subjective judgement of image quality, masking of films and time requirement were checked. In this investigation 68% of 175 Kodak EC-L ap/pa-films were judged "good", only 18% were classified "moderate" or "poor" 14%, but only 22% of 173 conventional ap/pa verification films (Agfa Curix HT 1000) were judged to be "good". The image quality, detail perception and time required for film inspection of the new Kodak EC-L film system was significantly improved when compared with standard portal films. They could be read more accurately and the detection of set-up deviation was facilitated.

  10. Effect of Immobilization and Performance Status on Intrafraction Motion for Stereotactic Lung Radiotherapy: Analysis of 133 Patients

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Li, Winnie, E-mail: winnie.li@rmp.uhn.on.ca; Department of Radiation Oncology, University of Toronto, Toronto, Ontario; Purdie, Thomas G.

    2011-12-01

    Purpose: To assess intrafractional geometric accuracy of lung stereotactic body radiation therapy (SBRT) patients treated with volumetric image guidance. Methods and Materials: Treatment setup accuracy was analyzed in 133 SBRT patients treated via research ethics board-approved protocols. For each fraction, a localization cone-beam computed tomography (CBCT) scan was acquired for soft-tissue registration to the internal target volume, followed by a couch adjustment for positional discrepancies greater than 3 mm, verified with a second CBCT scan. CBCT scans were also performed at intrafraction and end fraction. Patient positioning data from 2047 CBCT scans were recorded to determine systematic ({Sigma}) and randommore » ({sigma}) uncertainties, as well as planning target volume margins. Data were further stratified and analyzed by immobilization method (evacuated cushion [n = 75], evacuated cushion plus abdominal compression [n = 33], or chest board [n = 25]) and by patients' Eastern Cooperative Oncology Group performance status (PS): 0 (n = 31), 1 (n = 70), or 2 (n = 32). Results: Using CBCT internal target volume was matched within {+-}3 mm in 16% of all fractions at localization, 89% at verification, 72% during treatment, and 69% after treatment. Planning target volume margins required to encompass residual setup errors after couch corrections (verification CBCT scans) were 4 mm, and they increased to 5 mm with target intrafraction motion (post-treatment CBCT scans). Small differences (<1 mm) in the cranial-caudal direction of target position were observed between the immobilization cohorts in the localization, verification, intrafraction, and post-treatment CBCT scans (p < 0.01). Positional drift varied according to patient PS, with the PS 1 and 2 cohorts drifting out of position by mid treatment more than the PS 0 cohort in the cranial-caudal direction (p = 0.04). Conclusions: Image guidance ensures high geometric accuracy for lung SBRT irrespective of immobilization method or PS. A 5-mm setup margin suffices to address intrafraction motion. This setup margin may be further reduced by strategies such as frequent image guidance or volumetric arc therapy to correct or limit intrafraction motion.« less

  11. Effect of immobilization and performance status on intrafraction motion for stereotactic lung radiotherapy: analysis of 133 patients.

    PubMed

    Li, Winnie; Purdie, Thomas G; Taremi, Mojgan; Fung, Sharon; Brade, Anthony; Cho, B C John; Hope, Andrew; Sun, Alexander; Jaffray, David A; Bezjak, Andrea; Bissonnette, Jean-Pierre

    2011-12-01

    To assess intrafractional geometric accuracy of lung stereotactic body radiation therapy (SBRT) patients treated with volumetric image guidance. Treatment setup accuracy was analyzed in 133 SBRT patients treated via research ethics board-approved protocols. For each fraction, a localization cone-beam computed tomography (CBCT) scan was acquired for soft-tissue registration to the internal target volume, followed by a couch adjustment for positional discrepancies greater than 3 mm, verified with a second CBCT scan. CBCT scans were also performed at intrafraction and end fraction. Patient positioning data from 2047 CBCT scans were recorded to determine systematic (Σ) and random (σ) uncertainties, as well as planning target volume margins. Data were further stratified and analyzed by immobilization method (evacuated cushion [n=75], evacuated cushion plus abdominal compression [n=33], or chest board [n=25]) and by patients' Eastern Cooperative Oncology Group performance status (PS): 0 (n=31), 1 (n=70), or 2 (n=32). Using CBCT internal target volume was matched within ±3 mm in 16% of all fractions at localization, 89% at verification, 72% during treatment, and 69% after treatment. Planning target volume margins required to encompass residual setup errors after couch corrections (verification CBCT scans) were 4 mm, and they increased to 5 mm with target intrafraction motion (post-treatment CBCT scans). Small differences (<1 mm) in the cranial-caudal direction of target position were observed between the immobilization cohorts in the localization, verification, intrafraction, and post-treatment CBCT scans (p<0.01). Positional drift varied according to patient PS, with the PS 1 and 2 cohorts drifting out of position by mid treatment more than the PS 0 cohort in the cranial-caudal direction (p=0.04). Image guidance ensures high geometric accuracy for lung SBRT irrespective of immobilization method or PS. A 5-mm setup margin suffices to address intrafraction motion. This setup margin may be further reduced by strategies such as frequent image guidance or volumetric arc therapy to correct or limit intrafraction motion. Copyright © 2011 Elsevier Inc. All rights reserved.

  12. SU-F-P-23: Setup Uncertainties for the Lung Stereotactic Body Radiation Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zhang, Q; Vigneri, P; Madu, C

    2016-06-15

    Purpose: The Exactrack X-ray system with six degree-of-freedom (6DoF) adjustment ability can be used for setup of lung stereotactic body radiation therapy. The setup uncertainties from ExacTrack 6D system were analyzed. Methods: The Exactrack X-ray 6D image guided radiotherapy system is used in our clinic. The system is an integration of 2 subsystems: (1): an infrared based optical position system and (2) a radiography kV x-ray imaging system. The infrared system monitors reflective body markers on the patient’s skin to assistant in the initial setup. The radiographic kV devices were used for patient positions verification and adjustment. The position verificationmore » was made by fusing the radiographs with the digitally reconstructed radiograph (DRR) images generated by simulation CT images using 6DoF fusion algorithms. Those results were recorded in our system. Gaussian functions were used to fit the data. Results: For 37 lung SBRT patients, the image registration results for the initial setup by using surface markers and for the verifications, were measured. The results were analyzed for 143 treatments. The mean values for the lateral, longitudinal, vertical directions were 0.1, 0.3 and 0.3mm, respectively. The standard deviations for the lateral, longitudinal and vertical directions were 0.62, 0.78 and 0.75mm respectively. The mean values for the rotations around lateral, longitudinal and vertical directions were 0.1, 0.2 and 0.4 degrees respectively, with standard deviations of 0.36, 0.34, and 0.42 degrees. Conclusion: The setup uncertainties for the lung SBRT cases by using Exactrack 6D system were analyzed. The standard deviations of the setup errors were within 1mm for all three directions, and the standard deviations for rotations were within 0.5 degree.« less

  13. Initial experience in treating lung cancer with helical tomotherapy

    PubMed Central

    Yartsev, S; Dar, AR; Woodford, C; Wong, E; Bauman, G; Van Dyk, J

    2007-01-01

    Helical tomotherapy is a new form of image-guided radiation therapy that combines features of a linear accelerator and a helical computed tomography (CT) scanner. Megavoltage CT (MVCT) data allow the verification and correction of patient setup on the couch by comparison and image registration with the kilovoltage CT multi-slice images used for treatment planning. An 84-year-old male patient with Stage III bulky non-small cell lung cancer was treated on a Hi-ART II tomotherapy unit. Daily MVCT imaging was useful for setup corrections and signaled the need to adapt the delivery plan when the patient’s anatomy changed significantly. PMID:21614260

  14. Portal imaging based definition of the planning target volume during pelvic irradiation for gynecological malignancies.

    PubMed

    Mock, U; Dieckmann, K; Wolff, U; Knocke, T H; Pötter, R

    1999-08-01

    Geometrical accuracy in patient positioning can vary substantially during external radiotherapy. This study estimated the set-up accuracy during pelvic irradiation for gynecological malignancies for determination of safety margins (planning target volume, PTV). Based on electronic portal imaging devices (EPID), 25 patients undergoing 4-field pelvic irradiation for gynecological malignancies were analyzed with regard to set-up accuracy during the treatment course. Regularly performed EPID images were used in order to systematically assess the systematic and random component of set-up displacements. Anatomical matching of verification and simulation images was followed by measuring corresponding distances between the central axis and anatomical features. Data analysis of set-up errors referred to the x-, y-,and z-axes. Additionally, cumulative frequencies were evaluated. A total of 50 simulation films and 313 verification images were analyzed. For the anterior-posterior (AP) beam direction mean deviations along the x- and z-axes were 1.5 mm and -1.9 mm, respectively. Moreover, random errors of 4.8 mm (x-axis) and 3.0 mm (z-axis) were determined. Concerning the latero-lateral treatment fields, the systematic errors along the two axes were calculated to 2.9 mm (y-axis) and -2.0 mm (z-axis) and random errors of 3.8 mm and 3.5 mm were found, respectively. The cumulative frequency of misalignments < or =5 mm showed values of 75% (AP fields) and 72% (latero-lateral fields). With regard to cumulative frequencies < or =10 mm quantification revealed values of 97% for both beam directions. During external pelvic irradiation therapy for gynecological malignancies, EPID images on a regular basis revealed acceptable set-up inaccuracies. Safety margins (PTV) of 1 cm appear to be sufficient, accounting for more than 95% of all deviations.

  15. SU-F-J-24: Setup Uncertainty and Margin of the ExacTrac 6D Image Guide System for Patients with Brain Tumors

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kim, S; Oh, S; Yea, J

    Purpose: This study evaluated the setup uncertainties for brain sites when using BrainLAB’s ExacTrac X-ray 6D system for daily pretreatment to determine the optimal planning target volume (PTV) margin. Methods: Between August 2012 and April 2015, 28 patients with brain tumors were treated by daily image-guided radiotherapy using the BrainLAB ExacTrac 6D image guidance system of the Novalis-Tx linear accelerator. DUONTM (Orfit Industries, Wijnegem, Belgium) masks were used to fix the head. The radiotherapy was fractionated into 27–33 treatments. In total, 844 image verifications were performed for 28 patients and used for the analysis. The setup corrections along with themore » systematic and random errors were analyzed for six degrees of freedom in the translational (lateral, longitudinal, and vertical) and rotational (pitch, roll, and yaw) dimensions. Results: Optimal PTV margins were calculated based on van Herk et al.’s [margin recipe = 2.5∑ + 0.7σ − 3 mm] and Stroom et al.’s [margin recipe = 2∑ + 0.7σ] formulas. The systematic errors (∑) were 0.72, 1.57, and 0.97 mm in the lateral, longitudinal, and vertical translational dimensions, respectively, and 0.72°, 0.87°, and 0.83° in the pitch, roll, and yaw rotational dimensions, respectively. The random errors (σ) were 0.31, 0.46, and 0.54 mm in the lateral, longitudinal, and vertical rotational dimensions, respectively, and 0.28°, 0.24°, and 0.31° in the pitch, roll, and yaw rotational dimensions, respectively. According to van Herk et al.’s and Stroom et al.’s recipes, the recommended lateral PTV margins were 0.97 and 1.66 mm, respectively; the longitudinal margins were 1.26 and 3.47 mm, respectively; and the vertical margins were 0.21 and 2.31 mm, respectively. Conclusion: Therefore, daily setup verifications using the BrainLAB ExacTrac 6D image guide system are very useful for evaluating the setup uncertainties and determining the setup margin.∑σ.« less

  16. SU-F-J-32: Do We Need KV Imaging During CBCT Based Patient Set-Up for Lung Radiation Therapy?

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Gopal, A; Zhou, J; Prado, K

    Purpose: To evaluate the role of 2D kilovoltage (kV) imaging to complement cone beam CT (CBCT) imaging in a shift threshold based image guided radiation therapy (IGRT) strategy for conventional lung radiotherapy. Methods: A retrospective study was conducted by analyzing IGRT couch shift trends for 15 patients that received lung radiation therapy to evaluate the benefit of performing orthogonal kV imaging prior to CBCT imaging. Herein, a shift threshold based IGRT protocol was applied, which would mandate additional CBCT verification if the applied patient shifts exceeded 3 mm to avoid intraobserver variability in CBCT registration and to confirm table shifts.more » For each patient, two IGRT strategies: kV + CBCT and CBCT alone, were compared and the recorded patient shifts were categorized into whether additional CBCT acquisition would have been mandated or not. The effectiveness of either strategy was gauged by the likelihood of needing additional CBCT imaging for accurate patient set-up. Results: The use of CBCT alone was 6 times more likely to require an additional CBCT than KV+CBCT, for a 3 mm shift threshold (88% vs 14%). The likelihood of additional CBCT verification generally increased with lower shift thresholds, and was significantly lower when kV+CBCT was used (7% with 5 mm shift threshold, 36% with 2 mm threshold), than with CBCT alone (61% with 5 mm shift threshold, 97% with 2 mm threshold). With CBCT alone, treatment time increased by 2.2 min and dose increased by 1.9 cGy per fraction on average due to additional CBCT with a 3mm shift threshold. Conclusion: The benefit of kV imaging to screen for gross misalignments led to more accurate CBCT based patient localization compared with using CBCT alone. The subsequently reduced need for additional CBCT verification will minimize treatment time and result in less overall patient imaging dose.« less

  17. SU-E-T-287: Robustness Study of Passive-Scattering Proton Therapy in Lung: Is Range and Setup Uncertainty Calculation On the Initial CT Enough to Predict the Plan Robustness?

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ding, X; Dormer, J; Kenton, O

    Purpose: Plan robustness of the passive-scattering proton therapy treatment of lung tumors has been studied previously using combined uncertainties of 3.5% in CT number and 3 mm geometric shifts. In this study, we investigate whether this method is sufficient to predict proton plan robustness by comparing to plans performed on weekly verification CT scans. Methods: Ten lung cancer patients treated with passive-scattering proton therapy were randomly selected. All plans were prescribed 6660cGy in 37 fractions. Each initial plan was calculated using +/− 3.5% range and +/− 0.3cm setup uncertainty in x, y and z directions in Eclipse TPS(Method-A). Throughout themore » treatment course, patients received weekly verification CT scans to assess the daily treatment variation(Method-B). After contours and imaging registrations are verified by the physician, the initial plan with the same beamline and compensator was mapped into the verification CT. Dose volume histograms (DVH) were evaluated for robustness study. Results: Differences are observed between method A and B in terms of iCTV coverage and lung dose. Method-A shows all the iCTV D95 are within +/− 1% difference, while 20% of cases fall outside +/−1% range in Method-B. In the worst case scenario(WCS), the iCTV D95 is reduced by 2.5%. All lung V5 and V20 are within +/−5% in Method-A while 15% of V5 and 10% of V20 fall outside of +/−5% in Method-B. In the WCS, Lung V5 increased by 15% and V20 increased by 9%. Method A and B show good agreement with regard to cord maximum and Esophagus mean dose. Conclusion: This study suggests that using range and setup uncertainty calculation (+/−3.5% and +/−3mm) may not be sufficient to predict the WCS. In the absence of regular verification scans, expanding the conventional uncertainty parameters(e.g., to +/−3.5% and +/−4mm) may be needed to better reflect plan actual robustness.« less

  18. Proton therapy of prostate cancer by anterior-oblique beams: implications of setup and anatomy variations

    NASA Astrophysics Data System (ADS)

    Moteabbed, M.; Trofimov, A.; Sharp, G. C.; Wang, Y.; Zietman, A. L.; Efstathiou, J. A.; Lu, H.-M.

    2017-03-01

    Proton therapy of prostate by anterior beams could offer an attractive option for treating patients with hip prosthesis and limiting the high-dose exposure to the rectum. We investigated the impact of setup and anatomy variations on the anterior-oblique (AO) proton plan dose, and strategies to manage these effects via range verification and adaptive delivery. Ten patients treated by bilateral (BL) passive-scattering proton therapy (79.2 Gy in 44 fractions) who underwent weekly verification CT scans were selected. Plans with AO beams were additionally created. To isolate the effect of daily variations, initial AO plans did not include range uncertainty margins. The use of fixed planning margins and adaptive range adjustments to manage these effects was investigated. For each case, the planned dose was recalculated on weekly CTs, and accumulated on the simulation CT using deformable registration to approximate the delivered dose. Planned and accumulated doses were compared for each scenario to quantify dose deviations induced by variations. The possibility of estimating the necessary range adjustments before each treatment was explored by simulating the procedure of a diode-based in vivo range verification technique, which would potentially be used clinically. The average planned rectum, penile bulb and femoral heads mean doses were smaller for initial AO compared to BL plans (by 8.3, 16.1 and 25.9 Gy, respectively). After considering interfractional variations in AO plans, the target coverage was substantially reduced. The maximum reduction of V 79.2/D 95/D mean/EUD for AO (without distal margins) (25.3%/10.7/1.6/4.9 Gy, respectively) was considerably larger than BL plans. The loss of coverage was mainly related to changes in water equivalent path length of the prostate after fiducial-based setup, caused by discrepancies in patient anterior surface and bony-anatomy alignment. Target coverage was recovered partially when using fixed planning margins, and fully when applying adaptive range adjustments. The accumulated organs-at-risk dose for AO beams after range adjustment demonstrated full sparing of femoral heads and superior sparing of penile bulb and rectum compared to the conventional BL cases. Our study indicates that using AO beams makes prostate treatment more susceptible to target underdose induced by interfractional variations. Adaptive range verification/adjustment may facilitate the use of anterior beam approaches, and ensure adequate target coverage in every fraction of the treatment.

  19. Cone beam CT imaging with limited angle of projections and prior knowledge for volumetric verification of non-coplanar beam radiation therapy: a proof of concept study

    NASA Astrophysics Data System (ADS)

    Meng, Bowen; Xing, Lei; Han, Bin; Koong, Albert; Chang, Daniel; Cheng, Jason; Li, Ruijiang

    2013-11-01

    Non-coplanar beams are important for treatment of both cranial and noncranial tumors. Treatment verification of such beams with couch rotation/kicks, however, is challenging, particularly for the application of cone beam CT (CBCT). In this situation, only limited and unconventional imaging angles are feasible to avoid collision between the gantry, couch, patient, and on-board imaging system. The purpose of this work is to develop a CBCT verification strategy for patients undergoing non-coplanar radiation therapy. We propose an image reconstruction scheme that integrates a prior image constrained compressed sensing (PICCS) technique with image registration. Planning CT or CBCT acquired at the neutral position is rotated and translated according to the nominal couch rotation/translation to serve as the initial prior image. Here, the nominal couch movement is chosen to have a rotational error of 5° and translational error of 8 mm from the ground truth in one or more axes or directions. The proposed reconstruction scheme alternates between two major steps. First, an image is reconstructed using the PICCS technique implemented with total-variation minimization and simultaneous algebraic reconstruction. Second, the rotational/translational setup errors are corrected and the prior image is updated by applying rigid image registration between the reconstructed image and the previous prior image. The PICCS algorithm and rigid image registration are alternated iteratively until the registration results fall below a predetermined threshold. The proposed reconstruction algorithm is evaluated with an anthropomorphic digital phantom and physical head phantom. The proposed algorithm provides useful volumetric images for patient setup using projections with an angular range as small as 60°. It reduced the translational setup errors from 8 mm to generally <1 mm and the rotational setup errors from 5° to <1°. Compared with the PICCS algorithm alone, the integration of rigid registration significantly improved the reconstructed image quality, with a reduction of mostly 2-3 folds (up to 100) in root mean square image error. The proposed algorithm provides a remedy for solving the problem of non-coplanar CBCT reconstruction from limited angle of projections by combining the PICCS technique and rigid image registration in an iterative framework. In this proof of concept study, non-coplanar beams with couch rotations of 45° can be effectively verified with the CBCT technique.

  20. SU-F-T-642: Sub Millimeter Accurate Setup of More Than Three Vertebrae in Spinal SBRT with 6D Couch

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wang, X; Zhao, Z; Yang, J

    Purpose: To assess the initial setup accuracy in treating more than 3 vertebral body levels in spinal SBRT using a 6D couch. Methods: We retrospectively analyzed last 20 spinal SBRT patients (4 cervical, 9 thoracic, 7 lumbar/sacrum) treated in our clinic. These patients in customized immobilization device were treated in 1 or 3 fractions. Initial setup used ExacTrac and Brainlab 6D couch to align target within 1 mm and 1 degree, following by a cone beam CT (CBCT) for verification. Our current standard practice allows treating a maximum of three continuous vertebrae. Here we assess the possibility to achieve submore » millimeter setup accuracy for more than three vertebrae by examining the residual error in every slice of CBCT. The CBCT had a range of 17.5 cm, which covered 5 to 9 continuous vertebrae depending on the patient and target location. In the study, CBCT from the 1st fraction treatment was rigidly registered with the planning CT in Pinnacle. The residual setup error of a vertebra was determined by expanding the vertebra contour on the planning CT to be large enough to enclose the corresponding vertebra on CBCT. The margin of the expansion was considered as setup error. Results: Out of the 20 patients analyzed, initial setup accuracy can be achieved within 1 mm for a span of 5 or more vertebrae starting from T2 vertebra to inferior vertebra levels. 2 cervical and 2 upper thoracic patients showed the cervical spine was difficult to achieve sub millimeter accuracy for multi levels without a customized immobilization headrest. Conclusion: If the curvature of spinal columns can be reproduced in customized immobilization device during treatment as simulation, multiple continuous vertebrae can be setup within 1 mm with the use of a 6D couch.« less

  1. The role of a prone setup in breast radiation therapy.

    PubMed

    Huppert, Nelly; Jozsef, Gabor; Dewyngaert, Keith; Formenti, Silvia Chiara

    2011-01-01

    Most patients undergoing breast conservation therapy receive radiotherapy in the supine position. Historically, prone breast irradiation has been advocated for women with large pendulous breasts in order to decrease acute and late toxicities. With the advent of CT planning, the prone technique has become both feasible and reproducible. It was shown to be advantageous not only for women with larger breasts but in most patients since it consistently reduces, if not eliminates, the inclusion of heart and lung within the field. The prone setup has been accepted as the best localizing position for both MRI and stereotactic biopsy, but its adoption has been delayed in radiotherapy. New technological advances including image-modulated radiation therapy and image-guided radiation therapy have made possible the exploration of accelerated fractionation schemes with a concomitant boost to the tumor bed in the prone position, along with better imaging and verification of reproducibility of patient setup. This review describes some of the available techniques for prone breast radiotherapy and the available experience in their application. The NYU prone breast radiotherapy approach is discussed, including a summary of the results from several prospective trials.

  2. Initial clinical experience with a video-based patient positioning system.

    PubMed

    Johnson, L S; Milliken, B D; Hadley, S W; Pelizzari, C A; Haraf, D J; Chen, G T

    1999-08-01

    To report initial clinical experience with an interactive, video-based patient positioning system that is inexpensive, quick, accurate, and easy to use. System hardware includes two black-and-white CCD cameras, zoom lenses, and a PC equipped with a frame grabber. Custom software is used to acquire and archive video images, as well as to display real-time subtraction images revealing patient misalignment in multiple views. Two studies are described. In the first study, video is used to document the daily setup histories of 5 head and neck patients. Time-lapse cine loops are generated for each patient and used to diagnose and correct common setup errors. In the second study, 6 twice-daily (BID) head and neck patients are positioned according to the following protocol: at AM setups conventional treatment room lasers are used; at PM setups lasers are used initially and then video is used for 1-2 minutes to fine-tune the patient position. Lateral video images and lateral verification films are registered off-line to compare the distribution of setup errors per patient, with and without video assistance. In the first study, video images were used to determine the accuracy of our conventional head and neck setup technique, i.e., alignment of lightcast marks and surface anatomy to treatment room lasers and the light field. For this initial cohort of patients, errors ranged from sigma = 5 to 7 mm and were patient-specific. Time-lapse cine loops of the images revealed sources of the error, and as a result, our localization techniques and immobilization device were modified to improve setup accuracy. After the improvements, conventional setup errors were reduced to sigma = 3 to 5 mm. In the second study, when a stereo pair of live subtraction images were introduced to perform daily "on-line" setup correction, errors were reduced to sigma = 1 to 3 mm. Results depended on patient health and cooperation and the length of time spent fine-tuning the position. An interactive, video-based patient positioning system was shown to reduce setup errors to within 1 to 3 mm in head and neck patients, without a significant increase in overall treatment time or labor-intensive procedures. Unlike retrospective portal image analysis, use of two live-video images provides the therapists with immediate feedback and allows for true 3-D positioning and correction of out-of-plane rotation before radiation is delivered. With significant improvement in head and neck alignment and the elimination of setup errors greater than 3 to 5 mm, margins associated with treatment volumes potentially can be reduced, thereby decreasing normal tissue irradiation.

  3. Poster — Thur Eve — 16: 4DCT simulation with synchronized contrast injection of liver SBRT patients

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Karotki, A.; Korol, R.; Milot, L.

    2014-08-15

    Stereotactic body radiation therapy (SBRT) has recently emerged as a valid option for treating liver metastases. SBRT delivers highly conformai dose over a small number of fractions. As such it is particularly sensitive to the accuracy of target volume delineation by the radiation oncologist. However, contouring liver metastases remains challenging for the following reasons. First, the liver usually undergoes significant motion due to respiration. Second, liver metastases are often nearly indistinguishable from the surrounding tissue when using computed tomography (CT) for imaging making it difficult to identify and delineate them. Both problems can be overcome by using four dimensional CTmore » (4DCT) synchronized with intravenous contrast injection. We describe a novel CT simulation process which involves two 4DCT scans. The first scan captures the tumor and immediately surrounding tissue which in turn reduces the 4DCT scan time so that it can be optimally timed with intravenous contrast injection. The second 4DCT scan covers a larger volume and is used as the primary CT dataset for dose calculation, as well as patient setup verification on the treatment unit. The combination of two 4DCT scans, short and long, allows visualization of the liver metastases over all phases of breathing cycle while simultaneously acquiring long enough 4DCT dataset suitable for planning and patient setup verification.« less

  4. Cone-Beam Computed Tomography–Guided Positioning of Laryngeal Cancer Patients with Large Interfraction Time Trends in Setup and Nonrigid Anatomy Variations

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Gangsaas, Anne, E-mail: a.gangsaas@erasmusmc.nl; Astreinidou, Eleftheria; Quint, Sandra

    2013-10-01

    Purpose: To investigate interfraction setup variations of the primary tumor, elective nodes, and vertebrae in laryngeal cancer patients and to validate protocols for cone beam computed tomography (CBCT)-guided correction. Methods and Materials: For 30 patients, CBCT-measured displacements in fractionated treatments were used to investigate population setup errors and to simulate residual setup errors for the no action level (NAL) offline protocol, the extended NAL (eNAL) protocol, and daily CBCT acquisition with online analysis and repositioning. Results: Without corrections, 12 of 26 patients treated with radical radiation therapy would have experienced a gradual change (time trend) in primary tumor setup ≥4more » mm in the craniocaudal (CC) direction during the fractionated treatment (11/12 in caudal direction, maximum 11 mm). Due to these trends, correction of primary tumor displacements with NAL resulted in large residual CC errors (required margin 6.7 mm). With the weekly correction vector adjustments in eNAL, the trends could be largely compensated (CC margin 3.5 mm). Correlation between movements of the primary and nodal clinical target volumes (CTVs) in the CC direction was poor (r{sup 2}=0.15). Therefore, even with online setup corrections of the primary CTV, the required CC margin for the nodal CTV was as large as 6.8 mm. Also for the vertebrae, large time trends were observed for some patients. Because of poor CC correlation (r{sup 2}=0.19) between displacements of the primary CTV and the vertebrae, even with daily online repositioning of the vertebrae, the required CC margin around the primary CTV was 6.9 mm. Conclusions: Laryngeal cancer patients showed substantial interfraction setup variations, including large time trends, and poor CC correlation between primary tumor displacements and motion of the nodes and vertebrae (internal tumor motion). These trends and nonrigid anatomy variations have to be considered in the choice of setup verification protocol and planning target volume margins. eNAL could largely compensate time trends with minor prolongation of fraction time.« less

  5. Verification and implementation of set-up empirical models in pile design : research project capsule.

    DOT National Transportation Integrated Search

    2016-08-01

    The primary objectives of this research include: performing static and dynamic load tests on : newly instrumented test piles to better understand the set-up mechanism for individual soil : layers, verifying or recalibrating previously developed empir...

  6. An automatic dose verification system for adaptive radiotherapy for helical tomotherapy

    NASA Astrophysics Data System (ADS)

    Mo, Xiaohu; Chen, Mingli; Parnell, Donald; Olivera, Gustavo; Galmarini, Daniel; Lu, Weiguo

    2014-03-01

    Purpose: During a typical 5-7 week treatment of external beam radiotherapy, there are potential differences between planned patient's anatomy and positioning, such as patient weight loss, or treatment setup. The discrepancies between planned and delivered doses resulting from these differences could be significant, especially in IMRT where dose distributions tightly conforms to target volumes while avoiding organs-at-risk. We developed an automatic system to monitor delivered dose using daily imaging. Methods: For each treatment, a merged image is generated by registering the daily pre-treatment setup image and planning CT using treatment position information extracted from the Tomotherapy archive. The treatment dose is then computed on this merged image using our in-house convolution-superposition based dose calculator implemented on GPU. The deformation field between merged and planning CT is computed using the Morphon algorithm. The planning structures and treatment doses are subsequently warped for analysis and dose accumulation. All results are saved in DICOM format with private tags and organized in a database. Due to the overwhelming amount of information generated, a customizable tolerance system is used to flag potential treatment errors or significant anatomical changes. A web-based system and a DICOM-RT viewer were developed for reporting and reviewing the results. Results: More than 30 patients were analysed retrospectively. Our in-house dose calculator passed 97% gamma test evaluated with 2% dose difference and 2mm distance-to-agreement compared with Tomotherapy calculated dose, which is considered sufficient for adaptive radiotherapy purposes. Evaluation of the deformable registration through visual inspection showed acceptable and consistent results, except for cases with large or unrealistic deformation. Our automatic flagging system was able to catch significant patient setup errors or anatomical changes. Conclusions: We developed an automatic dose verification system that quantifies treatment doses, and provides necessary information for adaptive planning without impeding clinical workflows.

  7. Effect of electron contamination on in vivo dosimetry for lung block shielding during TBI

    PubMed Central

    Narayanasamy, Ganesh; Cruz, Wilbert; Saenz, Daniel L.; Stathakis, Sotirios; Papanikolaou, Niko

    2016-01-01

    Our institution performs in vivo verification measurement for each of our total body irradiation (TBI) patients with optically stimulated luminescent dosimeters (OSLD). The lung block verification measurements were commonly higher than expected. The aim of this work is to understand this discrepancy and improve the accuracy of these lung block verification measurements. Initially, the thickness of the lung block was increased to provide adequate lung sparing. Further tests revealed the increase was due to electron contamination dose emanating from the lung block. The thickness of the bolus material covering the OSLD behind the lung block was increased to offset the electron contamination. In addition, the distance from the lung block to the dosimeter was evaluated for its effect on the OSLD reading and found to be clinically insignificant over the range of variability in our clinic. The results show that the improved TBI treatment technique provides for better accuracy of measured dose in vivo and consistency of patient setup. PACS number(s): 87.53.Bn, 87.53.Kn, 87.55.N‐, 87.55.Qr PMID:27167290

  8. Prevention of gross setup errors in radiotherapy with an efficient automatic patient safety system.

    PubMed

    Yan, Guanghua; Mittauer, Kathryn; Huang, Yin; Lu, Bo; Liu, Chihray; Li, Jonathan G

    2013-11-04

    Treatment of the wrong body part due to incorrect setup is among the leading types of errors in radiotherapy. The purpose of this paper is to report an efficient automatic patient safety system (PSS) to prevent gross setup errors. The system consists of a pair of charge-coupled device (CCD) cameras mounted in treatment room, a single infrared reflective marker (IRRM) affixed on patient or immobilization device, and a set of in-house developed software. Patients are CT scanned with a CT BB placed over their surface close to intended treatment site. Coordinates of the CT BB relative to treatment isocenter are used as reference for tracking. The CT BB is replaced with an IRRM before treatment starts. PSS evaluates setup accuracy by comparing real-time IRRM position with reference position. To automate system workflow, PSS synchronizes with the record-and-verify (R&V) system in real time and automatically loads in reference data for patient under treatment. Special IRRMs, which can permanently stick to patient face mask or body mold throughout the course of treatment, were designed to minimize therapist's workload. Accuracy of the system was examined on an anthropomorphic phantom with a designed end-to-end test. Its performance was also evaluated on head and neck as well as abdominalpelvic patients using cone-beam CT (CBCT) as standard. The PSS system achieved a seamless clinic workflow by synchronizing with the R&V system. By permanently mounting specially designed IRRMs on patient immobilization devices, therapist intervention is eliminated or minimized. Overall results showed that the PSS system has sufficient accuracy to catch gross setup errors greater than 1 cm in real time. An efficient automatic PSS with sufficient accuracy has been developed to prevent gross setup errors in radiotherapy. The system can be applied to all treatment sites for independent positioning verification. It can be an ideal complement to complex image-guidance systems due to its advantages of continuous tracking ability, no radiation dose, and fully automated clinic workflow.

  9. A Noninvasive Body Setup Method for Radiotherapy by Using a Multimodal Image Fusion Technique

    PubMed Central

    Zhang, Jie; Chen, Yunxia; Wang, Chenchen; Chu, Kaiyue; Jin, Jianhua; Huang, Xiaolin; Guan, Yue; Li, Weifeng

    2017-01-01

    Purpose: To minimize the mismatch error between patient surface and immobilization system for tumor location by a noninvasive patient setup method. Materials and Methods: The method, based on a point set registration, proposes a shift for patient positioning by integrating information of the computed tomography scans and that of optical surface landmarks. An evaluation of the method included 3 areas: (1) a validation on a phantom by estimating 100 known mismatch errors between patient surface and immobilization system. (2) Five patients with pelvic tumors were considered. The tumor location errors of the method were measured using the difference between the proposal shift of cone-beam computed tomography and that of our method. (3) The collected setup data from the evaluation of patients were compared with the published performance data of other 2 similar systems. Results: The phantom verification results showed that the method was capable of estimating mismatch error between patient surface and immobilization system in a precision of <0.22 mm. For the pelvic tumor, the method had an average tumor location error of 1.303, 2.602, and 1.684 mm in left–right, anterior–posterior, and superior–inferior directions, respectively. The performance comparison with other 2 similar systems suggested that the method had a better positioning accuracy for pelvic tumor location. Conclusion: By effectively decreasing an interfraction uncertainty source (mismatch error between patient surface and immobilization system) in radiotherapy, the method can improve patient positioning precision for pelvic tumor. PMID:29333959

  10. Clinical application of in vivo treatment delivery verification based on PET/CT imaging of positron activity induced at high energy photon therapy

    NASA Astrophysics Data System (ADS)

    Janek Strååt, Sara; Andreassen, Björn; Jonsson, Cathrine; Noz, Marilyn E.; Maguire, Gerald Q., Jr.; Näfstadius, Peder; Näslund, Ingemar; Schoenahl, Frederic; Brahme, Anders

    2013-08-01

    The purpose of this study was to investigate in vivo verification of radiation treatment with high energy photon beams using PET/CT to image the induced positron activity. The measurements of the positron activation induced in a preoperative rectal cancer patient and a prostate cancer patient following 50 MV photon treatments are presented. A total dose of 5 and 8 Gy, respectively, were delivered to the tumors. Imaging was performed with a 64-slice PET/CT scanner for 30 min, starting 7 min after the end of the treatment. The CT volume from the PET/CT and the treatment planning CT were coregistered by matching anatomical reference points in the patient. The treatment delivery was imaged in vivo based on the distribution of the induced positron emitters produced by photonuclear reactions in tissue mapped on to the associated dose distribution of the treatment plan. The results showed that spatial distribution of induced activity in both patients agreed well with the delivered beam portals of the treatment plans in the entrance subcutaneous fat regions but less so in blood and oxygen rich soft tissues. For the preoperative rectal cancer patient however, a 2 ± (0.5) cm misalignment was observed in the cranial-caudal direction of the patient between the induced activity distribution and treatment plan, indicating a beam patient setup error. No misalignment of this kind was seen in the prostate cancer patient. However, due to a fast patient setup error in the PET/CT scanner a slight mis-position of the patient in the PET/CT was observed in all three planes, resulting in a deformed activity distribution compared to the treatment plan. The present study indicates that the induced positron emitters by high energy photon beams can be measured quite accurately using PET imaging of subcutaneous fat to allow portal verification of the delivered treatment beams. Measurement of the induced activity in the patient 7 min after receiving 5 Gy involved count rates which were about 20 times lower than that of a patient undergoing standard 18F-FDG treatment. When using a combination of short lived nuclides such as 15O (half-life: 2 min) and 11C (half-life: 20 min) with low activity it is not optimal to use clinical reconstruction protocols. Thus, it might be desirable to further optimize reconstruction parameters as well as to address hardware improvements in realizing in vivo treatment verification with PET/CT in the future. A significant improvement with regard to 15O imaging could also be expected by having the PET/CT unit located close to the radiation treatment room.

  11. Tailoring four-dimensional cone-beam CT acquisition settings for fiducial marker-based image guidance in radiation therapy.

    PubMed

    Jin, Peng; van Wieringen, Niek; Hulshof, Maarten C C M; Bel, Arjan; Alderliesten, Tanja

    2018-04-01

    Use of four-dimensional cone-beam CT (4D-CBCT) and fiducial markers for image guidance during radiation therapy (RT) of mobile tumors is challenging due to the trade-off among image quality, imaging dose, and scanning time. This study aimed to investigate different 4D-CBCT acquisition settings for good visibility of fiducial markers in 4D-CBCT. Using these 4D-CBCTs, the feasibility of marker-based 4D registration for RT setup verification and manual respiration-induced motion quantification was investigated. For this, we applied a dynamic phantom with three different breathing motion amplitudes and included two patients with implanted markers. Irrespective of the motion amplitude, for a medium field of view (FOV), marker visibility was improved by reducing the imaging dose per projection and increasing the number of projection images; however, the scanning time was 4 to 8 min. For a small FOV, the total imaging dose and the scanning time were reduced (62.5% of the dose using a medium FOV, 2.5 min) without losing marker visibility. However, the body contour could be missing for a small FOV, which is not preferred in RT. The marker-based 4D setup verification was feasible for both the phantom and patient data. Moreover, manual marker motion quantification can achieve a high accuracy with a mean error of [Formula: see text].

  12. SU-E-J-16: Automatic Image Contrast Enhancement Based On Automatic Parameter Optimization for Radiation Therapy Setup Verification

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Qiu, J; Washington University in St Louis, St Louis, MO; Li, H. Harlod

    Purpose: In RT patient setup 2D images, tissues often cannot be seen well due to the lack of image contrast. Contrast enhancement features provided by image reviewing software, e.g. Mosaiq and ARIA, require manual selection of the image processing filters and parameters thus inefficient and cannot be automated. In this work, we developed a novel method to automatically enhance the 2D RT image contrast to allow automatic verification of patient daily setups as a prerequisite step of automatic patient safety assurance. Methods: The new method is based on contrast limited adaptive histogram equalization (CLAHE) and high-pass filtering algorithms. The mostmore » important innovation is to automatically select the optimal parameters by optimizing the image contrast. The image processing procedure includes the following steps: 1) background and noise removal, 2) hi-pass filtering by subtracting the Gaussian smoothed Result, and 3) histogram equalization using CLAHE algorithm. Three parameters were determined through an iterative optimization which was based on the interior-point constrained optimization algorithm: the Gaussian smoothing weighting factor, the CLAHE algorithm block size and clip limiting parameters. The goal of the optimization is to maximize the entropy of the processed Result. Results: A total 42 RT images were processed. The results were visually evaluated by RT physicians and physicists. About 48% of the images processed by the new method were ranked as excellent. In comparison, only 29% and 18% of the images processed by the basic CLAHE algorithm and by the basic window level adjustment process, were ranked as excellent. Conclusion: This new image contrast enhancement method is robust and automatic, and is able to significantly outperform the basic CLAHE algorithm and the manual window-level adjustment process that are currently used in clinical 2D image review software tools.« less

  13. SU-E-J-15: Automatically Detect Patient Treatment Position and Orientation in KV Portal Images

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Qiu, J; Yang, D

    2015-06-15

    Purpose: In the course of radiation therapy, the complex information processing workflow will Result in potential errors, such as incorrect or inaccurate patient setups. With automatic image check and patient identification, such errors could be effectively reduced. For this purpose, we developed a simple and rapid image processing method, to automatically detect the patient position and orientation in 2D portal images, so to allow automatic check of positions and orientations for patient daily RT treatments. Methods: Based on the principle of portal image formation, a set of whole body DRR images were reconstructed from multiple whole body CT volume datasets,more » and fused together to be used as the matching template. To identify the patient setup position and orientation shown in a 2D portal image, the 2D portal image was preprocessed (contrast enhancement, down-sampling and couch table detection), then matched to the template image so to identify the laterality (left or right), position, orientation and treatment site. Results: Five day’s clinical qualified portal images were gathered randomly, then were processed by the automatic detection and matching method without any additional information. The detection results were visually checked by physicists. 182 images were correct detection in a total of 200kV portal images. The correct rate was 91%. Conclusion: The proposed method can detect patient setup and orientation quickly and automatically. It only requires the image intensity information in KV portal images. This method can be useful in the framework of Electronic Chart Check (ECCK) to reduce the potential errors in workflow of radiation therapy and so to improve patient safety. In addition, the auto-detection results, as the patient treatment site position and patient orientation, could be useful to guide the sequential image processing procedures, e.g. verification of patient daily setup accuracy. This work was partially supported by research grant from Varian Medical System.« less

  14. Practical Use of the Extended No Action Level (eNAL) Correction Protocol for Breast Cancer Patients With Implanted Surgical Clips

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Penninkhof, Joan, E-mail: j.penninkhof@erasmusmc.nl; Quint, Sandra; Baaijens, Margreet

    Purpose: To describe the practical use of the extended No Action Level (eNAL) setup correction protocol for breast cancer patients with surgical clips and evaluate its impact on the setup accuracy of both tumor bed and whole breast during simultaneously integrated boost treatments. Methods and Materials: For 80 patients, two orthogonal planar kilovoltage images and one megavoltage image (for the mediolateral beam) were acquired per fraction throughout the radiotherapy course. For setup correction, the eNAL protocol was applied, based on registration of surgical clips in the lumpectomy cavity. Differences with respect to application of a No Action Level (NAL) protocolmore » or no protocol were quantified for tumor bed and whole breast. The correlation between clip migration during the fractionated treatment and either the method of surgery or the time elapsed from last surgery was investigated. Results: The distance of the clips to their center of mass (COM), averaged over all clips and patients, was reduced by 0.9 {+-} 1.2 mm (mean {+-} 1 SD). Clip migration was similar between the group of patients starting treatment within 100 days after surgery (median, 53 days) and the group starting afterward (median, 163 days) (p = 0.20). Clip migration after conventional breast surgery (closing the breast superficially) or after lumpectomy with partial breast reconstructive techniques (sutured cavity). was not significantly different either (p = 0.22). Application of eNAL on clips resulted in residual systematic errors for the clips' COM of less than 1 mm in each direction, whereas the setup of the breast was within about 2 mm of accuracy. Conclusions: Surgical clips can be safely used for high-accuracy position verification and correction. Given compensation for time trends in the clips' COM throughout the treatment course, eNAL resulted in better setup accuracies for both tumor bed and whole breast than NAL.« less

  15. Accuracy Evaluation of a 3-Dimensional Surface Imaging System for Guidance in Deep-Inspiration Breath-Hold Radiation Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Alderliesten, Tanja; Sonke, Jan-Jakob; Betgen, Anja

    2013-02-01

    Purpose: To investigate the applicability of 3-dimensional (3D) surface imaging for image guidance in deep-inspiration breath-hold radiation therapy (DIBH-RT) for patients with left-sided breast cancer. For this purpose, setup data based on captured 3D surfaces was compared with setup data based on cone beam computed tomography (CBCT). Methods and Materials: Twenty patients treated with DIBH-RT after breast-conserving surgery (BCS) were included. Before the start of treatment, each patient underwent a breath-hold CT scan for planning purposes. During treatment, dose delivery was preceded by setup verification using CBCT of the left breast. 3D surfaces were captured by a surface imaging systemmore » concurrently with the CBCT scan. Retrospectively, surface registrations were performed for CBCT to CT and for a captured 3D surface to CT. The resulting setup errors were compared with linear regression analysis. For the differences between setup errors, group mean, systematic error, random error, and 95% limits of agreement were calculated. Furthermore, receiver operating characteristic (ROC) analysis was performed. Results: Good correlation between setup errors was found: R{sup 2}=0.70, 0.90, 0.82 in left-right, craniocaudal, and anterior-posterior directions, respectively. Systematic errors were {<=}0.17 cm in all directions. Random errors were {<=}0.15 cm. The limits of agreement were -0.34-0.48, -0.42-0.39, and -0.52-0.23 cm in left-right, craniocaudal, and anterior-posterior directions, respectively. ROC analysis showed that a threshold between 0.4 and 0.8 cm corresponds to promising true positive rates (0.78-0.95) and false positive rates (0.12-0.28). Conclusions: The results support the application of 3D surface imaging for image guidance in DIBH-RT after BCS.« less

  16. Simulation of Dose to Surrounding Normal Structures in Tangential Breast Radiotherapy Due to Setup Error

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Prabhakar, Ramachandran; Department of Nuclear Medicine, All India Institute of Medical Sciences, New Delhi; Department of Radiology, All India Institute of Medical Sciences, New Delhi

    Setup error plays a significant role in the final treatment outcome in radiotherapy. The effect of setup error on the planning target volume (PTV) and surrounding critical structures has been studied and the maximum allowed tolerance in setup error with minimal complications to the surrounding critical structure and acceptable tumor control probability is determined. Twelve patients were selected for this study after breast conservation surgery, wherein 8 patients were right-sided and 4 were left-sided breast. Tangential fields were placed on the 3-dimensional-computed tomography (3D-CT) dataset by isocentric technique and the dose to the PTV, ipsilateral lung (IL), contralateral lung (CLL),more » contralateral breast (CLB), heart, and liver were then computed from dose-volume histograms (DVHs). The planning isocenter was shifted for 3 and 10 mm in all 3 directions (X, Y, Z) to simulate the setup error encountered during treatment. Dosimetric studies were performed for each patient for PTV according to ICRU 50 guidelines: mean doses to PTV, IL, CLL, heart, CLB, liver, and percentage of lung volume that received a dose of 20 Gy or more (V20); percentage of heart volume that received a dose of 30 Gy or more (V30); and volume of liver that received a dose of 50 Gy or more (V50) were calculated for all of the above-mentioned isocenter shifts and compared to the results with zero isocenter shift. Simulation of different isocenter shifts in all 3 directions showed that the isocentric shifts along the posterior direction had a very significant effect on the dose to the heart, IL, CLL, and CLB, which was followed by the lateral direction. The setup error in isocenter should be strictly kept below 3 mm. The study shows that isocenter verification in the case of tangential fields should be performed to reduce future complications to adjacent normal tissues.« less

  17. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Horst, Astrid van der, E-mail: a.vanderhorst@amc.uva.nl; Lens, Eelco; Wognum, Silvia

    Purpose: Because of low soft-tissue contrast of cone beam computed tomography (CBCT), fiducial markers are often used for radiation therapy patient setup verification. For pancreatic cancer patients, biliary stents have been suggested as surrogate fiducials. Using intratumoral fiducials as standard for tumor position, this study aims to quantify the suitability of biliary stents for measuring interfractional and respiratory-induced position variations of pancreatic tumors. Methods and Materials: Eleven pancreatic cancer patients with intratumoral fiducials and a biliary stent were included in this study. Daily CBCT scans (243 in total) were registered with a reference CT scan, based on bony anatomy, onmore » fiducial markers, and on the biliary stent, respectively. We analyzed the differences in tumor position (ie, markers center-of-mass position) among these 3 registrations. In addition, we measured for 9 patients the magnitude of respiratory-induced motion (MM) of the markers and of the stent on 4-dimensional CT (4DCT) and determined the difference between these 2 magnitudes (ΔMM). Results: The stent indicated tumor position better than bony anatomy in 67% of fractions; the absolute difference between the markers and stent registration was >5 mm in 46% of fractions and >10 mm in 20% of fractions. Large PTV margins (superior-inferior direction, >19 mm) would be needed to account for this interfractional position variability. On 4DCT, we found in superior-inferior direction a mean ΔMM of 0.5 mm (range, –2.6 to 4.2 mm). Conclusions: For respiratory-induced motion, the mean ΔMM is small, but for individual patients the absolute difference can be >4 mm. For interfractional position variations, a stent is, on average, a better surrogate fiducial than bony anatomy, but large PTV margins would still be required. Therefore, intratumoral fiducials are recommended for online setup verification for all pancreatic patients scheduled for radiation therapy, including patients with a biliary stent.« less

  18. Spine stereotactic body radiotherapy utilizing cone-beam CT image-guidance with a robotic couch: intrafraction motion analysis accounting for all six degrees of freedom.

    PubMed

    Hyde, Derek; Lochray, Fiona; Korol, Renee; Davidson, Melanie; Wong, C Shun; Ma, Lijun; Sahgal, Arjun

    2012-03-01

    To evaluate the residual setup error and intrafraction motion following kilovoltage cone-beam CT (CBCT) image guidance, for immobilized spine stereotactic body radiotherapy (SBRT) patients, with positioning corrected for in all six degrees of freedom. Analysis is based on 42 consecutive patients (48 thoracic and/or lumbar metastases) treated with a total of 106 fractions and 307 image registrations. Following initial setup, a CBCT was acquired for patient alignment and a pretreatment CBCT taken to verify shifts and determine the residual setup error, followed by a midtreatment and posttreatment CBCT image. For 13 single-fraction SBRT patients, two midtreatment CBCT images were obtained. Initially, a 1.5-mm and 1° tolerance was used to reposition the patient following couch shifts which was subsequently reduced to 1 mm and 1° degree after the first 10 patients. Small positioning errors after the initial CBCT setup were observed, with 90% occurring within 1 mm and 97% within 1°. In analyzing the impact of the time interval for verification imaging (10 ± 3 min) and subsequent image acquisitions (17 ± 4 min), the residual setup error was not significantly different (p > 0.05). A significant difference (p = 0.04) in the average three-dimensional intrafraction positional deviations favoring a more strict tolerance in translation (1 mm vs. 1.5 mm) was observed. The absolute intrafraction motion averaged over all patients and all directions along x, y, and z axis (± SD) were 0.7 ± 0.5 mm and 0.5 ± 0.4 mm for the 1.5 mm and 1 mm tolerance, respectively. Based on a 1-mm and 1° correction threshold, the target was localized to within 1.2 mm and 0.9° with 95% confidence. Near-rigid body immobilization, intrafraction CBCT imaging approximately every 15-20 min, and strict repositioning thresholds in six degrees of freedom yields minimal intrafraction motion allowing for safe spine SBRT delivery. Copyright © 2012 Elsevier Inc. All rights reserved.

  19. SU-F-J-130: Margin Determination for Hypofractionated Partial Breast Irradiation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Geady, C; Keller, B; Hahn, E

    2016-06-15

    Purpose: To determine the Planning Target Volume (PTV) margin for Hypofractionated Partial Breast Irradiation (HPBI) using the van Herk formalism (M=2.5∑+0.7σ). HPBI is a novel technique intended to provide local control in breast cancer patients not eligible for surgical resection, using 40 Gy in 5 fractions prescribed to the gross disease. Methods: Setup uncertainties were quantified through retrospective analysis of cone-beam computed tomography (CBCT) data sets, collected prior to (prefraction) and after (postfraction) treatment delivery. During simulation and treatment, patients were immobilized using a wing board and an evacuated bag. Prefraction CBCT was rigidly registered to planning 4-dimensional computed tomographymore » (4DCT) using the chest wall and tumor, and translational couch shifts were applied as needed. This clinical workflow was faithfully reproduced in Pinnacle (Philips Medical Systems) to yield residual setup and intrafractional error through translational shifts and rigid registrations (ribs and sternum) of prefraction CBCT to 4DCT and postfraction CBCT to prefraction CBCT, respectively. All ten patients included in this investigation were medically inoperable; the median age was 84 (range, 52–100) years. Results: Systematic (and random) setup uncertainties (in mm) detected for the left-right, craniocaudal and anteroposterior directions were 0.4 (1.5), 0.8 (1.8) and 0.4 (1.0); net uncertainty was determined to be 0.7 (1.5). Rotations >2° in any axis occurred on 8/72 (11.1%) registrations. Conclusion: Preliminary results suggest a non-uniform setup margin (in mm) of 2.2, 3.3 and 1.7 for the left-right, craniocaudal and anteroposterior directions is required for HPBI, given its immobilization techniques and online setup verification protocol. This investigation is ongoing, though published results from similar studies are consistent with the above findings. Determination of margins in breast radiotherapy is a paradigm shift, but a necessary step in moving towards hypofractionated regiments, which may ultimately redefine the standard of care for this select patient population.« less

  20. SCA security verification on wireless sensor network node

    NASA Astrophysics Data System (ADS)

    He, Wei; Pizarro, Carlos; de la Torre, Eduardo; Portilla, Jorge; Riesgo, Teresa

    2011-05-01

    Side Channel Attack (SCA) differs from traditional mathematic attacks. It gets around of the exhaustive mathematic calculation and precisely pin to certain points in the cryptographic algorithm to reveal confidential information from the running crypto-devices. Since the introduction of SCA by Paul Kocher et al [1], it has been considered to be one of the most critical threats to the resource restricted but security demanding applications, such as wireless sensor networks. In this paper, we focus our work on the SCA-concerned security verification on WSN (wireless sensor network). A detailed setup of the platform and an analysis of the results of DPA (power attack) and EMA (electromagnetic attack) is presented. The setup follows the way of low-cost setup to make effective SCAs. Meanwhile, surveying the weaknesses of WSNs in resisting SCA attacks, especially for the EM attack. Finally, SCA-Prevention suggestions based on Differential Security Strategy for the FPGA hardware implementation in WSN will be given, helping to get an improved compromise between security and cost.

  1. Ultrasound functional imaging in an ex vivo beating porcine heart platform

    NASA Astrophysics Data System (ADS)

    Petterson, Niels J.; Fixsen, Louis S.; Rutten, Marcel C. M.; Pijls, Nico H. J.; van de Vosse, Frans N.; Lopata, Richard G. P.

    2017-12-01

    In recent years, novel ultrasound functional imaging (UFI) techniques have been introduced to assess cardiac function by measuring, e.g. cardiac output (CO) and/or myocardial strain. Verification and reproducibility assessment in a realistic setting remain major issues. Simulations and phantoms are often unrealistic, whereas in vivo measurements often lack crucial hemodynamic parameters or ground truth data, or suffer from the large physiological and clinical variation between patients when attempting clinical validation. Controlled validation in certain pathologies is cumbersome and often requires the use of lab animals. In this study, an isolated beating pig heart setup was adapted and used for performance assessment of UFI techniques such as volume assessment and ultrasound strain imaging. The potential of performing verification and reproducibility studies was demonstrated. For proof-of-principle, validation of UFI in pathological hearts was examined. Ex vivo porcine hearts (n  =  6, slaughterhouse waste) were resuscitated and attached to a mock circulatory system. Radio frequency ultrasound data of the left ventricle were acquired in five short axis views and one long axis view. Based on these slices, the CO was measured, where verification was performed using flow sensor measurements in the aorta. Strain imaging was performed providing radial, circumferential and longitudinal strain to assess reproducibility and inter-subject variability under steady conditions. Finally, strains in healthy hearts were compared to a heart with an implanted left ventricular assist device, simulating a failing, supported heart. Good agreement between ultrasound and flow sensor based CO measurements was found. Strains were highly reproducible (intraclass correlation coefficients  >0.8). Differences were found due to biological variation and condition of the hearts. Strain magnitude and patterns in the assisted heart were available for different pump action, revealing large changes compared to the normal condition. The setup provides a valuable benchmarking platform for UFI techniques. Future studies will include work on different pathologies and other means of measurement verification.

  2. A comparative study between the imaging system and the optical tracking system in proton therapy at CNAO

    PubMed Central

    Desplanques, Maxime; Tagaste, Barbara; Fontana, Giulia; Pella, Andrea; Riboldi, Marco; Fattori, Giovanni; Donno, Andrea; Baroni, Guido; Orecchia, Roberto

    2013-01-01

    The synergy between in-room imaging and optical tracking, in co-operation with highly accurate robotic patient handling represents a concept for patient-set-up which has been implemented at CNAO (Centro Nazionale di Adroterapia Oncologica). In-room imaging is based on a double oblique X-ray projection system; optical tracking consists of the detection of the position of spherical markers placed directly on the patient's skin or on the immobilization devices. These markers are used as external fiducials during patient positioning and dose delivery. This study reports the results of a comparative analysis between in-room imaging and optical tracking data for patient positioning within the framework of high-precision particle therapy. Differences between the optical tracking system (OTS) and the imaging system (IS) were on average within the expected localization accuracy. On the first 633 fractions for head and neck (H&N) set-up procedures, the corrections applied by the IS, after patient positioning using the OTS only, were for the mostly sub-millimetric regarding the translations (0.4±1.1 mm) and sub-gradual regarding the rotations (0.0°±0.8°). On the first 236 fractions for pelvis localizations the amplitude of the corrections applied by the IS after preliminary optical set-up correction were moderately higher and more dispersed (translations: 1.3±2.9 mm, rotations 0.1±0.9°). Although the indication of the OTS cannot replace information provided by in-room imaging devices and 2D-3D image registration, the reported data show that OTS preliminary correction might greatly support image-based patient set-up refinement and also provide a secondary, independent verification system for patient positioning. PMID:23824116

  3. SU-E-CAMPUS-J-05: Quantitative Investigation of Random and Systematic Uncertainties From Hardware and Software Components in the Frameless 6DBrainLAB ExacTrac System

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Keeling, V; Jin, H; Hossain, S

    2014-06-15

    Purpose: To evaluate setup accuracy and quantify individual systematic and random errors for the various hardware and software components of the frameless 6D-BrainLAB ExacTrac system. Methods: 35 patients with cranial lesions, some with multiple isocenters (50 total lesions treated in 1, 3, 5 fractions), were investigated. All patients were simulated with a rigid head-and-neck mask and the BrainLAB localizer. CT images were transferred to the IPLAN treatment planning system where optimized plans were generated using stereotactic reference frame based on the localizer. The patients were setup initially with infrared (IR) positioning ExacTrac system. Stereoscopic X-ray images (XC: X-ray Correction) weremore » registered to their corresponding digitally-reconstructed-radiographs, based on bony anatomy matching, to calculate 6D-translational and rotational (Lateral, Longitudinal, Vertical, Pitch, Roll, Yaw) shifts. XC combines systematic errors of the mask, localizer, image registration, frame, and IR. If shifts were below tolerance (0.7 mm translational and 1 degree rotational), treatment was initiated; otherwise corrections were applied and additional X-rays were acquired to verify patient position (XV: X-ray Verification). Statistical analysis was used to extract systematic and random errors of the different components of the 6D-ExacTrac system and evaluate the cumulative setup accuracy. Results: Mask systematic errors (translational; rotational) were the largest and varied from one patient to another in the range (−15 to 4mm; −2.5 to 2.5degree) obtained from mean of XC for each patient. Setup uncertainty in IR positioning (0.97,2.47,1.62mm;0.65,0.84,0.96degree) was extracted from standard-deviation of XC. Combined systematic errors of the frame and localizer (0.32,−0.42,−1.21mm; −0.27,0.34,0.26degree) was extracted from mean of means of XC distributions. Final patient setup uncertainty was obtained from the standard deviations of XV (0.57,0.77,0.67mm,0.39,0.35,0.30degree). Conclusion: Statistical analysis was used to calculate cumulative and individual systematic errors from the different hardware and software components of the 6D-ExacTrac-system. Patients were treated with cumulative errors (<1mm,<1degree) with XV image guidance.« less

  4. Ground vibration tests of a high fidelity truss for verification of on orbit damage location techniques

    NASA Technical Reports Server (NTRS)

    Kashangaki, Thomas A. L.

    1992-01-01

    This paper describes a series of modal tests that were performed on a cantilevered truss structure. The goal of the tests was to assemble a large database of high quality modal test data for use in verification of proposed methods for on orbit model verification and damage detection in flexible truss structures. A description of the hardware is provided along with details of the experimental setup and procedures for 16 damage cases. Results from selected cases are presented and discussed. Differences between ground vibration testing and on orbit modal testing are also described.

  5. Personalized monitoring of therapeutic salicylic acid in dried blood spots using a three-layer setup and desorption electrospray ionization mass spectrometry.

    PubMed

    Siebenhaar, Markus; Küllmer, Kai; Fernandes, Nuno Miguel de Barros; Hüllen, Volker; Hopf, Carsten

    2015-09-01

    Desorption electrospray ionization (DESI) mass spectrometry is an emerging technology for direct therapeutic drug monitoring in dried blood spots (DBS). Current DBS methods require manual application of small molecules as internal standards for absolute drug quantification. With industrial standardization in mind, we superseded the manual addition of standard and built a three-layer setup for robust quantification of salicylic acid directly from DBS. We combined a dioctyl sodium sulfosuccinate weave facilitating sample spreading with a cellulose layer for addition of isotope-labeled salicylic acid as internal standard and a filter paper for analysis of the standard-containing sample by DESI-MS. Using this setup, we developed a quantification method for salicylic acid from whole blood with a validated linear curve range from 10 to 2000 mg/L, a relative standard deviation (RSD%) ≤14%, and determination coefficients of 0.997. The limit of detection (LOD) was 8 mg/L and the lower limit of quantification (LLOQ) was 10 mg/L. Recovery rates in method verification by LC-MS/MS were 97 to 101% for blinded samples. Most importantly, a study in healthy volunteers after administration of a single dose of Aspirin provides evidence to suggest that the three-layer setup may enable individual pharmacokinetic and endpoint testing following blood collection by finger pricking by patients at home. Taken together, our data suggests that DBS-based quantification of drugs by DESI-MS on pre-manufactured three-layer cartridges may be a promising approach for future near-patient therapeutic drug monitoring.

  6. Experimental setup for the measurement of induction motor cage currents

    NASA Astrophysics Data System (ADS)

    Bottauscio, Oriano; Chiampi, Mario; Donadio, Lorenzo; Zucca, Mauro

    2005-04-01

    An experimental setup for measurement of the currents flowing in the rotor bars of induction motors during synchronous no-load tests is described in the paper. The experimental verification of the high-frequency phenomena in the rotor cage is fundamental for a deep insight of the additional loss estimation by numerical methods. The attention is mainly focused on the analysis and design of the transducers developed for the cage current measurement.

  7. Prospective study of cone-beam computed tomography image-guided radiotherapy for prone accelerated partial breast irradiation.

    PubMed

    Jozsef, Gabor; DeWyngaert, J Keith; Becker, Stewart J; Lymberis, Stella; Formenti, Silvia C

    2011-10-01

    To report setup variations during prone accelerated partial breast irradiation (APBI). New York University (NYU) 07-582 is an institutional review board-approved protocol of cone-beam computed tomography (CBCT) to deliver image-guided ABPI in the prone position. Eligible are postmenopausal women with pT1 breast cancer excised with negative margins and no nodal involvement. A total dose of 30 Gy in five daily fractions of 6 Gy are delivered to the planning target volume (the tumor cavity with 1.5-cm margin) by image-guided radiotherapy. Patients are set up prone, on a dedicated mattress, used for both simulation and treatment. After positioning with skin marks and lasers, CBCTs are performed and the images are registered to the planning CT. The resulting shifts (setup corrections) are recorded in the three principal directions and applied. Portal images are taken for verification. If they differ from the planning digital reconstructed radiographs, the patient is reset, and a new CBCT is taken. 70 consecutive patients have undergone a total of 343 CBCTs: 7 patients had four of five planned CBCTs performed. Seven CBCTs (2%) required to be repeated because of misalignment in the comparison between portal and digital reconstructed radiograph image after the first CBCT. The mean shifts and standard deviations in the anterior-posterior (AP), superior-inferior (SI), and medial-lateral (ML) directions were -0.19 (0.54), -0.02 (0.33), and -0.02 (0.43) cm, respectively. The average root mean squares of the daily shifts were 0.50 (0.28), 0.29 (0.17), and 0.38 (0.20). A conservative margin formula resulted in a recommended margin of 1.26, 0.73, 0.96 cm in the AP, SI, and ML directions. CBCTs confirmed that the NYU prone APBI setup and treatment technique are reproducible, with interfraction variation comparable to those reported for supine setup. The currently applied margin (1.5 cm) adequately compensates for the setup variation detected. Copyright © 2011 Elsevier Inc. All rights reserved.

  8. 4D cone-beam CT imaging for guidance in radiation therapy: setup verification by use of implanted fiducial markers

    NASA Astrophysics Data System (ADS)

    Jin, Peng; van Wieringen, Niek; Hulshof, Maarten C. C. M.; Bel, Arjan; Alderliesten, Tanja

    2016-03-01

    The use of 4D cone-beam computed tomography (CBCT) and fiducial markers for guidance during radiation therapy of mobile tumors is challenging due to the trade-off between image quality, imaging dose, and scanning time. We aimed to investigate the visibility of markers and the feasibility of marker-based 4D registration and manual respiration-induced marker motion quantification for different CBCT acquisition settings. A dynamic thorax phantom and a patient with implanted gold markers were included. For both the phantom and patient, the peak-to-peak amplitude of marker motion in the cranial-caudal direction ranged from 5.3 to 14.0 mm, which did not affect the marker visibility and the associated marker-based registration feasibility. While using a medium field of view (FOV) and the same total imaging dose as is applied for 3D CBCT scanning in our clinic, it was feasible to attain an improved marker visibility by reducing the imaging dose per projection and increasing the number of projection images. For a small FOV with a shorter rotation arc but similar total imaging dose, streak artifacts were reduced due to using a smaller sampling angle. Additionally, the use of a small FOV allowed reducing total imaging dose and scanning time (~2.5 min) without losing the marker visibility. In conclusion, by using 4D CBCT with identical or lower imaging dose and a reduced gantry speed, it is feasible to attain sufficient marker visibility for marker-based 4D setup verification. Moreover, regardless of the settings, manual marker motion quantification can achieve a high accuracy with the error <1.2 mm.

  9. SU-F-T-287: A Preliminary Study On Patient Specific VMAT Verification Using a Phosphor-Screen Based Geometric QA System (Raven QA)

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lee, M; Yi, B; Wong, J

    Purpose: The RavenQA system (LAP Laser, Germany) is a QA device with a phosphor screen detector for performing the QA tasks of TG-142. This study tested if it is feasible to use the system for the patient specific QA of the Volumetric Modulated Arc Therapy (VMAT). Methods: Water equivalent material (5cm) is attached to the front of the detector plate of the RavenQA for dosimetry purpose. Then the plate is attached to the gantry to synchronize the movement between the detector and the gantry. Since the detector moves together with gantry, The ’Reset gantry to 0’ function of the Eclipsemore » planning system (Varian, CA) is used to simulate the measurement situation when calculating dose of the detector plate. The same gantry setup is used when delivering the treatment beam for feasibility test purposes. Cumulative dose is acquired for each arc. The optical scatter component of each captured image from the CCD camera is corrected by deconvolving the 2D spatial invariant optical scatter kernel (OSK). We assume that the OSK is a 2D isotropic point spread function with inverse-squared decrease as a function of radius from the center. Results: Three cases of VMAT plans including head & neck, whole pelvis and abdomen-pelvis are tested. Setup time for measurements was less than 5 minutes. Passing rates of absolute gamma were 99.3, 98.2, 95.9 respectively for 3%/3mm criteria and 96.2, 97.1, 86.4 for 2%/2mm criteria. The abdomen-pelvis field has long treatment fields, 37cm, which are longer than the detector plate (25cm). This plan showed relatively lower passing rate than other plans. Conclusion: An algorithm for IMRT/VMAT verification using the RavenQA has been developed and tested. The model of spatially invariant OSK works well for deconvolution purpose. It is proved that the RavenQA can be used for the patient specific verification of VMAT. This work is funded in part by a Maryland Industrial Partnership Program grant to University of Maryland and to JPLC who owns the Raven technology. John Wong is a co-founder of JPLC.« less

  10. Setup, Validation and Quality Control of a Centralized WGS laboratory - Lessons Learned.

    PubMed

    Arnold, Cath; Edwards, Kirstin; Desai, Meeta; Platt, Steve; Green, Jonathan; Conway, David

    2018-04-25

    Routine use of Whole Genome analysis for infectious diseases can be used to enlighten various scenarios pertaining to Public Health, including identification of microbial pathogens; relating individual cases to an outbreak of infectious disease; establishing an association between an outbreak of food poisoning and a specific food vehicle; inferring drug susceptibility; source tracing of contaminants and study of variations in the genome affect pathogenicity/virulence. We describe the setup, validation and ongoing verification of a centralised WGS laboratory to carry out the sequencing for these public health functions for the National Infection Services, Public Health England in the UK. The performance characteristics and Quality Control metrics measured during validation and verification of the entire end to end process (accuracy, precision, reproducibility and repeatability) are described and include information regarding the automated pass and release of data to service users without intervention. © Crown copyright 2018.

  11. Improved setup and positioning accuracy using a three‐point customized cushion/mask/bite‐block immobilization system for stereotactic reirradiation of head and neck cancer

    PubMed Central

    Wang, He; Wang, Congjun; Tung, Samuel; Dimmitt, Andrew Wilson; Wong, Pei Fong; Edson, Mark A.; Garden, Adam S.; Rosenthal, David I.; Fuller, Clifton D.; Gunn, Gary B.; Takiar, Vinita; Wang, Xin A.; Luo, Dershan; Yang, James N.; Wong, Jennifer

    2016-01-01

    The purpose of this study was to investigate the setup and positioning uncertainty of a custom cushion/mask/bite‐block (CMB) immobilization system and determine PTV margin for image‐guided head and neck stereotactic ablative radiotherapy (HN‐SABR). We analyzed 105 treatment sessions among 21 patients treated with HN‐SABR for recurrent head and neck cancers using a custom CMB immobilization system. Initial patient setup was performed using the ExacTrac infrared (IR) tracking system and initial setup errors were based on comparison of ExacTrac IR tracking system to corrected online ExacTrac X‐rays images registered to treatment plans. Residual setup errors were determined using repeat verification X‐ray. The online ExacTrac corrections were compared to cone‐beam CT (CBCT) before treatment to assess agreement. Intrafractional positioning errors were determined using prebeam X‐rays. The systematic and random errors were analyzed. The initial translational setup errors were −0.8±1.3 mm, −0.8±1.6 mm, and 0.3±1.9 mm in AP, CC, and LR directions, respectively, with a three‐dimensional (3D) vector of 2.7±1.4 mm. The initial rotational errors were up to 2.4° if 6D couch is not available. CBCT agreed with ExacTrac X‐ray images to within 2 mm and 2.5°. The intrafractional uncertainties were 0.1±0.6 mm, 0.1±0.6 mm, and 0.2±0.5 mm in AP, CC, and LR directions, respectively, and 0.0∘±0.5°, 0.0∘±0.6°, and −0.1∘±0.4∘ in yaw, roll, and pitch direction, respectively. The translational vector was 0.9±0.6 mm. The calculated PTV margins mPTV(90,95) were within 1.6 mm when using image guidance for online setup correction. The use of image guidance for online setup correction, in combination with our customized CMB device, highly restricted target motion during treatments and provided robust immobilization to ensure minimum dose of 95% to target volume with 2.0 mm PTV margin for HN‐SABR. PACS number(s): 87.55.ne PMID:27167275

  12. Implications of Artefacts Reduction in the Planning CT Originating from Implanted Fiducial Markers

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kassim, Iskandar, E-mail: i.binkassim@erasmusmc.n; Joosten, Hans; Barnhoorn, Jaco C.

    The efficacy of metal artefact reduction (MAR) software to suppress artefacts in reconstructed computed tomography (CT) images originating from small metal objects, like tumor markers and surgical clips, was evaluated. In addition, possible implications of using digital reconstructed radiographs (DRRs), based on the MAR CT images, for setup verification were analyzed. A phantom and 15 patients with different tumor sites and implanted markers were imaged with a multislice CT scanner. The raw image data was reconstructed both with the clinically used filtered-backprojection (FBP) and with the MAR software. Using the MAR software, improvements in image quality were often observed inmore » CT slices with markers or clips. Especially when several markers were located near to each other, fewer streak artefacts were observed than with the FBP algorithm. In addition, the shape and size of markers could be identified more accurately, reducing the contoured marker volumes by a factor of 2. For the phantom study, the CT numbers measured near to the markers corresponded more closely to the expected values. However, the MAR images were slightly more smoothed compared with the images reconstructed with FBP. For 8 prostate cancer patients in this study, the interobserver variation in 3D marker definition was similar (<0.4 mm) when using DRRs based on either FBP or MAR CT scans. Automatic marker matches also showed a similar success rate. However, differences in automatic match results up to 1 mm, caused by differences in the marker definition, were observed, which turned out to be (borderline) statistically significant (p = 0.06) for 2 patients. In conclusion, the MAR software might improve image quality by suppressing metal artefacts, probably allowing for a more reliable delineation of structures. When implanted markers or clips are used for setup verification, the accuracy may slightly be improved as well, which is relevant when using very tight clinical target volume (CTV) to planning target volume (PTV) margins for planning.« less

  13. MO-G-BRE-04: Automatic Verification of Daily Treatment Deliveries and Generation of Daily Treatment Reports for a MR Image-Guided Treatment Machine

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Yang, D; Li, X; Li, H

    2014-06-15

    Purpose: Two aims of this work were to develop a method to automatically verify treatment delivery accuracy immediately after patient treatment and to develop a comprehensive daily treatment report to provide all required information for daily MR-IGRT review. Methods: After systematically analyzing the requirements for treatment delivery verification and understanding the available information from a novel MR-IGRT treatment machine, we designed a method to use 1) treatment plan files, 2) delivery log files, and 3) dosimetric calibration information to verify the accuracy and completeness of daily treatment deliveries. The method verifies the correctness of delivered treatment plans and beams, beammore » segments, and for each segment, the beam-on time and MLC leaf positions. Composite primary fluence maps are calculated from the MLC leaf positions and the beam-on time. Error statistics are calculated on the fluence difference maps between the plan and the delivery. We also designed the daily treatment delivery report by including all required information for MR-IGRT and physics weekly review - the plan and treatment fraction information, dose verification information, daily patient setup screen captures, and the treatment delivery verification results. Results: The parameters in the log files (e.g. MLC positions) were independently verified and deemed accurate and trustable. A computer program was developed to implement the automatic delivery verification and daily report generation. The program was tested and clinically commissioned with sufficient IMRT and 3D treatment delivery data. The final version has been integrated into a commercial MR-IGRT treatment delivery system. Conclusion: A method was developed to automatically verify MR-IGRT treatment deliveries and generate daily treatment reports. Already in clinical use since December 2013, the system is able to facilitate delivery error detection, and expedite physician daily IGRT review and physicist weekly chart review.« less

  14. Evaluations of the setup discrepancy between BrainLAB 6D ExacTrac and cone-beam computed tomography used with the imaging guidance system Novalis-Tx for intracranial stereotactic radiosurgery.

    PubMed

    Oh, Se An; Park, Jae Won; Yea, Ji Woon; Kim, Sung Kyu

    2017-01-01

    The objective of this study was to evaluate the setup discrepancy between BrainLAB 6 degree-of-freedom (6D) ExacTrac and cone-beam computed tomography (CBCT) used with the imaging guidance system Novalis Tx for intracranial stereotactic radiosurgery. We included 107 consecutive patients for whom white stereotactic head frame masks (R408; Clarity Medical Products, Newark, OH) were used to fix the head during intracranial stereotactic radiosurgery, between August 2012 and July 2016. The patients were immobilized in the same state for both the verification image using 6D ExacTrac and online 3D CBCT. In addition, after radiation treatment, registration between the computed tomography simulation images and the CBCT images was performed with offline 6D fusion in an offline review. The root-mean-square of the difference in the translational dimensions between the ExacTrac system and CBCT was <1.01 mm for online matching and <1.10 mm for offline matching. Furthermore, the root-mean-square of the difference in the rotational dimensions between the ExacTrac system and the CBCT were <0.82° for online matching and <0.95° for offline matching. It was concluded that while the discrepancies in residual setup errors between the ExacTrac 6D X-ray and the CBCT were minor, they should not be ignored.

  15. [Verification of the dose delivered to the patient by means of TLD, SC, PID. What future?].

    PubMed

    Noël, A

    2003-11-01

    Among the different possibilities to check the accuracy of the treatment delivered, only in vivo dosimetry ensures the precision of the dose delivered to the patient during the treatment. In 1970-1980, Ruden assessed the use of thermoluminescent dosimetry to perform in vivo measurements at Radiumemmet in Stockholm. Straightforward in its principle but demanding in its implementation, thermoluminescent dosimetry has largely been used. Today, thanks to the work of Rikner, the use of semiconductor detectors allows the general implementation of in vivo dosimetry. Tomorrow, we will use electronic portal imaging device to verify the geometrical patient setup and the dose delivery at the same time. Its implementation remains complex and will need the development of algorithms to compute exit dose or midplane dose using portal in vivo dosimetry. First clinical results show that portal imaging is an accurate alternative for conventional in vivo dosimetry using diodes.

  16. Verification of the quantum dimension effects in electricsl condactivity with different topology of laser-induced thin-film structures

    NASA Astrophysics Data System (ADS)

    Arakelian, S.; Kucherik, A.; Kutrovskaya, S.; Osipov, A.; Istratov, A.; Skryabin, I.

    2018-01-01

    A clear physical model for the quantum states verification in nanocluster structures with jump/tunneling electroconductivity are under study in both theory and experiment. The accent is made on consideration of low-dimensional structures when the structural phase transitions occur and the tendency to high enhancement electroconductivity obtained. The results give us an opportunity to establish a basis for new physical principles to create the functional elements for the optoelectronics and photonics in hybrid set-up (optics + electrophysics) by the nanocluster technology approach.

  17. ExacTrac Snap Verification: A New Tool for Ensuring Quality Control for Lung Stereotactic Body Radiation Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Udrescu, Corina; Mornex, Francoise, E-mail: francoise.mornex@chu-lyon.fr; Tanguy, Ronan

    2013-01-01

    Purpose: The intrafraction verification provided by ExacTrac X-ray 6D Snap Verification (ET-SV) allows the tracking of potential isocenter displacements throughout patient position and treatment. The aims of this study were (1) to measure the intrafraction variations of the isocenter position (random errors); (2) to study the amplitude of the variation related to the fraction duration; and (3) to assess the impact of the table movement on positioning uncertainties. Methods and Materials: ET-SV uses images acquired before or during treatment delivery or both to detect isocenter displacement. Twenty patients treated with stereotactic body radiation therapy (SBRT) for lung tumors underwent SVmore » before or during each beam. Noncoplanar beams were sometimes necessary. The time between the setup of the patient and each SV was noted, and values of deviations were compiled for 3 SV time groups: SV performed at {<=}10 min (group 1), between 11 and 20 min (group 2), and {>=}21 min (group 3). Random errors in positioning during the use of noncoplanar fields were noted. Results: The mean isocenter deviation {+-}SD was 2 {+-} 0.5 mm (range, 1-8 mm). The average deviations {+-}SD increased significantly from 1.6 {+-} 0.5 mm to 2.1 {+-} 0.8 mm and 2.2 {+-} 0.6 mm for groups 1, 2, and 3 (P=.002), respectively. Percentages of deviation {>=}3 mm were 7.06%, 22.83%, and 28.07% and 1.08%, 4.15%, and 8.4% for {>=}5 mm (P<.0001). For 11 patients, table rotation was necessary. The mean isocenter deviation {+-}SD increased significantly from 1.9 {+-} 0.5 mm before table rotation to 2.7 {+-} 0.5 mm (P=.001) for the first beam treated after rotation. Conclusions: SV detects isocenter deviations, which increase in amplitude and frequency with the fraction duration, and enables intrafraction verification for SBRT (taking into account clinical condition and technical issues). SV gives accurate targeting at any time during irradiation and may raise confidence to escalate the dose. SV appears to be an important tool for ensuring the quality control of SBRT.« less

  18. Using refraction in thick glass plates for optical path length modulation in low coherence interferometry.

    PubMed

    Kröger, Niklas; Schlobohm, Jochen; Pösch, Andreas; Reithmeier, Eduard

    2017-09-01

    In Michelson interferometer setups the standard way to generate different optical path lengths between a measurement arm and a reference arm relies on expensive high precision linear stages such as piezo actuators. We present an alternative approach based on the refraction of light at optical interfaces using a cheap stepper motor with high gearing ratio to control the rotation of a glass plate. The beam path is examined and a relation between angle of rotation and change in optical path length is devised. As verification, an experimental setup is presented, and reconstruction results from a measurement standard are shown. The reconstructed step height from this setup lies within 1.25% of the expected value.

  19. Comparison of carina-based versus bony anatomy-based registration for setup verification in esophageal cancer radiotherapy.

    PubMed

    Machiels, Mélanie; Jin, Peng; van Gurp, Christianne H; van Hooft, Jeanin E; Alderliesten, Tanja; Hulshof, Maarten C C M

    2018-03-21

    To investigate the feasibility and geometric accuracy of carina-based registration for CBCT-guided setup verification in esophageal cancer IGRT, compared with current practice bony anatomy-based registration. Included were 24 esophageal cancer patients with 65 implanted fiducial markers, visible on planning CTs and follow-up CBCTs. All available CBCT scans (n = 236) were rigidly registered to the planning CT with respect to the bony anatomy and the carina. Target coverage was visually inspected and marker position variation was quantified relative to both registration approaches; the variation of systematic (Σ) and random errors (σ) was estimated. Automatic carina-based registration was feasible in 94.9% of the CBCT scans, with an adequate target coverage in 91.1% compared to 100% after bony anatomy-based registration. Overall, Σ (σ) in the LR/CC/AP direction was 2.9(2.4)/4.1(2.4)/2.2(1.8) mm using the bony anatomy registration compared to 3.3(3.0)/3.6(2.6)/3.9(3.1) mm for the carina. Mid-thoracic placed markers showed a non-significant but smaller Σ in CC and AP direction when using the carina-based registration. Compared with a bony anatomy-based registration, carina-based registration for esophageal cancer IGRT results in inadequate target coverage in 8.9% of cases. Furthermore, large Σ and σ, requiring larger anisotropic margins, were seen after carina-based registration. Only for tumors entirely confined to the mid-thoracic region the carina-based registration might be slightly favorable.

  20. Inexpensive Eddy-Current Standard

    NASA Technical Reports Server (NTRS)

    Berry, Robert F., Jr.

    1985-01-01

    Radial crack replicas serve as evaluation standards. Technique entails intimately joining two pieces of appropriate aluminum alloy stock and centering drilled hole through and along interface. Bore surface of hole presents two vertical stock interface lines 180 degrees apart. These lines serve as radial crack defect replicas during eddy-current technique setup and verification.

  1. SU-E-T-132: Dosimetric Impact of Positioning Errors in Hypo-Fractionated Cranial Radiation Therapy Using Frameless Stereotactic BrainLAB System

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Keeling, V; Jin, H; Ali, I

    2014-06-01

    Purpose: To determine dosimetric impact of positioning errors in the stereotactic hypo-fractionated treatment of intracranial lesions using 3Dtransaltional and 3D-rotational corrections (6D) frameless BrainLAB ExacTrac X-Ray system. Methods: 20 cranial lesions, treated in 3 or 5 fractions, were selected. An infrared (IR) optical positioning system was employed for initial patient setup followed by stereoscopic kV X-ray radiographs for position verification. 6D-translational and rotational shifts were determined to correct patient position. If these shifts were above tolerance (0.7 mm translational and 1° rotational), corrections were applied and another set of X-rays was taken to verify patient position. Dosimetric impact (D95, Dmin,more » Dmax, and Dmean of planning target volume (PTV) compared to original plans) of positioning errors for initial IR setup (XC: Xray Correction) and post-correction (XV: X-ray Verification) was determined in a treatment planning system using a method proposed by Yue et al. (Med. Phys. 33, 21-31 (2006)) with 3D-translational errors only and 6D-translational and rotational errors. Results: Absolute mean translational errors (±standard deviation) for total 92 fractions (XC/XV) were 0.79±0.88/0.19±0.15 mm (lateral), 1.66±1.71/0.18 ±0.16 mm (longitudinal), 1.95±1.18/0.15±0.14 mm (vertical) and rotational errors were 0.61±0.47/0.17±0.15° (pitch), 0.55±0.49/0.16±0.24° (roll), and 0.68±0.73/0.16±0.15° (yaw). The average changes (loss of coverage) in D95, Dmin, Dmax, and Dmean were 4.5±7.3/0.1±0.2%, 17.8±22.5/1.1±2.5%, 0.4±1.4/0.1±0.3%, and 0.9±1.7/0.0±0.1% using 6Dshifts and 3.1±5.5/0.0±0.1%, 14.2±20.3/0.8±1.7%, 0.0±1.2/0.1±0.3%, and 0.7±1.4/0.0±0.1% using 3D-translational shifts only. The setup corrections (XC-XV) improved the PTV coverage by 4.4±7.3% (D95) and 16.7±23.5% (Dmin) using 6D adjustment. Strong correlations were observed between translation errors and deviations in dose coverage for XC. Conclusion: The initial BrainLAB IR system based on rigidity of the mask-frame setup is not sufficient for accurate stereotactic positioning; however, with X-ray imageguidance sub-millimeter accuracy is achieved with negligible deviations in dose coverage. The angular corrections (mean angle summation=1.84°) are important and cause considerable deviations in dose coverage.« less

  2. Sierra/Aria 4.48 Verification Manual.

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sierra Thermal Fluid Development Team

    Presented in this document is a portion of the tests that exist in the Sierra Thermal/Fluids verification test suite. Each of these tests is run nightly with the Sierra/TF code suite and the results of the test checked under mesh refinement against the correct analytic result. For each of the tests presented in this document the test setup, derivation of the analytic solution, and comparison of the code results to the analytic solution is provided. This document can be used to confirm that a given code capability is verified or referenced as a compilation of example problems.

  3. Postmastectomy radiotherapy with integrated scar boost using helical tomotherapy.

    PubMed

    Rong, Yi; Yadav, Poonam; Welsh, James S; Fahner, Tasha; Paliwal, Bhudatt

    2012-01-01

    The purpose of this study was to evaluate helical tomotherapy dosimetry in postmastectomy patients undergoing treatment for chest wall and positive nodal regions with simultaneous integrated boost (SIB) in the scar region using strip bolus. Six postmastectomy patients were scanned with a 5-mm-thick strip bolus covering the scar planning target volume (PTV) plus 2-cm margin. For all 6 cases, the chest wall received a total cumulative dose of 49.3-50.4 Gy with daily fraction size of 1.7-2.0 Gy. Total dose to the scar PTV was prescribed to 58.0-60.2 Gy at 2.0-2.5 Gy per fraction. The supraclavicular PTV and mammary nodal PTV received 1.7-1.9 dose per fraction. Two plans (with and without bolus) were generated for all 6 cases. To generate no-bolus plans, strip bolus was contoured and overrode to air density before planning. The setup reproducibility and delivered dose accuracy were evaluated for all 6 cases. Dose-volume histograms were used to evaluate dose-volume coverage of targets and critical structures. We observed reduced air cavities with the strip bolus setup compared with what we normally see with the full bolus. The thermoluminescence dosimeters (TLD) in vivo dosimetry confirmed accurate dose delivery beneath the bolus. The verification plans performed on the first day megavoltage computed tomography (MVCT) image verified that the daily setup and overall dose delivery was within 2% accuracy compared with the planned dose. The hotspot of the scar PTV in no-bolus plans was 111.4% of the prescribed dose averaged over 6 cases compared with 106.6% with strip bolus. With a strip bolus only covering the postmastectomy scar region, we observed increased dose uniformity to the scar PTV, higher setup reproducibility, and accurate dose delivered beneath the bolus. This study demonstrates the feasibility of using a strip bolus over the scar using tomotherapy for SIB dosimetry in postmastectomy treatments. Published by Elsevier Inc.

  4. Postmastectomy radiotherapy with integrated scar boost using helical tomotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rong Yi, E-mail: rong@humonc.wisc.edu; University of Wisconsin Riverview Cancer Center, Wisconsin Rapids, WI; Yadav, Poonam

    2012-10-01

    The purpose of this study was to evaluate helical tomotherapy dosimetry in postmastectomy patients undergoing treatment for chest wall and positive nodal regions with simultaneous integrated boost (SIB) in the scar region using strip bolus. Six postmastectomy patients were scanned with a 5-mm-thick strip bolus covering the scar planning target volume (PTV) plus 2-cm margin. For all 6 cases, the chest wall received a total cumulative dose of 49.3-50.4 Gy with daily fraction size of 1.7-2.0 Gy. Total dose to the scar PTV was prescribed to 58.0-60.2 Gy at 2.0-2.5 Gy per fraction. The supraclavicular PTV and mammary nodal PTVmore » received 1.7-1.9 dose per fraction. Two plans (with and without bolus) were generated for all 6 cases. To generate no-bolus plans, strip bolus was contoured and overrode to air density before planning. The setup reproducibility and delivered dose accuracy were evaluated for all 6 cases. Dose-volume histograms were used to evaluate dose-volume coverage of targets and critical structures. We observed reduced air cavities with the strip bolus setup compared with what we normally see with the full bolus. The thermoluminescence dosimeters (TLD) in vivo dosimetry confirmed accurate dose delivery beneath the bolus. The verification plans performed on the first day megavoltage computed tomography (MVCT) image verified that the daily setup and overall dose delivery was within 2% accuracy compared with the planned dose. The hotspot of the scar PTV in no-bolus plans was 111.4% of the prescribed dose averaged over 6 cases compared with 106.6% with strip bolus. With a strip bolus only covering the postmastectomy scar region, we observed increased dose uniformity to the scar PTV, higher setup reproducibility, and accurate dose delivered beneath the bolus. This study demonstrates the feasibility of using a strip bolus over the scar using tomotherapy for SIB dosimetry in postmastectomy treatments.« less

  5. SU-E-T-64: A Programmable Moving Insert for the ArcCHECK Phantom for Dose Verification of Respiratory-Gated VMAT

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Gaede, S; Jordan, K; Western University, London, ON

    Purpose: To present a customized programmable moving insert for the ArcCHECK™ phantom that can, in a single delivery, check both entrance dosimetry, while simultaneously verifying the delivery of respiratory-gated VMAT. Methods: The cylindrical motion phantom uses a computer-controlled stepping motor to move an insert inside a stationery sleeve. Insert motion is programmable and can include rotational motion in addition to linear motion along the axis of the cylinder. The sleeve fits securely in the bore of the ArcCHECK™. Interchangeable inserts, including an A1SL chamber, optically-stimulated luminescence dosimeters, radiochromic film, or 3D gels, allow this combination to be used for commissioning,more » routine quality assurance, and patient-specific dosimetric verification of respiratory-gated VMAT. Before clinical implementation, the effect of a moving insert on the ArcCHECK™ measurements was considered. First, the measured dose to the ArcCHECK™ containing multiple inserts in the static position was compared to the calculated dose during multiple VMAT treatment deliveries. Then, dose was measured under both sinusoidal and real-patient motion conditions to determine any effect of the moving inserts on the ArcCHECK™ measurements. Finally, dose was measured during gated VMAT delivery to the same inserts under the same motion conditions to examine any effect of various beam “on-and-off” and dose rate ramp “up-and-down”. Multiple comparisons between measured and calculated dose to different inserts were also considered. Results: The pass rate for the static delivery exceeded 98% for all measurements (3%/3mm), suggesting a valid setup for entrance dosimetry. The pass rate was not altered for any measurement delivered under motion conditions. A similar Result was observed under gated VMAT conditions, including agreement of measured and calculated dose to the various inserts. Conclusion: Incorporating a programmable moving insert within the ArcCHECK™ phantom provides an efficient verification of respiratory-gated VMAT delivery that is useful during commissioning, routine quality assurance, and patient-specific dose verification. Prototype phantom development and testing was performed in collaboration with Modus Medical Devices Inc. (London, ON). No financial support was granted.« less

  6. TH-A-9A-05: Initial Setup Accuracy Comparison Between Frame-Based and Frameless Stereotactic Radiosurgery

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Tseng, T; Sheu, R; Todorov, B

    2014-06-15

    Purpose: To evaluate initial setup accuracy for stereotactic radiosurgery (SRS) between Brainlab frame-based and frameless immobilization system, also to discern the magnitude frameless system has on setup parameters. Methods: The correction shifts from the original setup were compared for total 157 SRS cranial treatments (69 frame-based vs. 88 frameless). All treatments were performed on a Novalis linac with ExacTrac positioning system. Localization box with isocenter overlay was used for initial setup and correction shift was determined by ExacTrac 6D auto-fusion to achieve submillimeter accuracy for treatment. For frameless treatments, mean time interval between simulation and treatment was 5.7 days (rangemore » 0–13). Pearson Chi-Square was used for univariate analysis. Results: The correctional radial shifts (mean±STD, median) for the frame and frameless system measured by ExacTrac were 1.2±1.2mm, 1.1mm and 3.1±3.3mm, 2.0mm, respectively. Treatments with frameless system had a radial shift >2mm more often than those with frames (51.1% vs. 2.9%; p<.0001). To achieve submillimeter accuracy, 85.5% frame-based treatments did not require shift and only 23.9% frameless treatment could succeed with initial setup. There was no statistical significant system offset observed in any direction for either system. For frameless treatments, those treated ≥ 3 days from simulation had statistically higher rates of radial shifts between 1–2mm and >2mm compared to patients treated in a shorter amount of time from simulation (34.3% and 56.7% vs. 28.6% and 33.3%, respectively; p=0.006). Conclusion: Although image-guided positioning system can also achieve submillimeter accuracy for frameless system, users should be cautious regarding the inherent uncertainty of its capability of immobilization. A proper quality assurance procedure for frameless mask manufacturing and a protocol for intra-fraction imaging verification will be crucial for frameless system. Time interval between simulation and treatment was influential to initial setup accuracy. A shorter time frame for frameless SRS treatment could be helpful in minimizing uncertainties in localization.« less

  7. SU-F-T-463: Light-Field Based Dynalog Verification

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Atwal, P; Ramaseshan, R

    2016-06-15

    Purpose: To independently verify leaf positions in so-called dynalog files for a Varian iX linac with a Millennium 120 MLC. This verification provides a measure of confidence that the files can be used directly as part of a more extensive intensity modulated radiation therapy / volumetric modulated arc therapy QA program. Methods: Initial testing used white paper placed at the collimator plane and a standard hand-held digital camera to image the light and shadow of a static MLC field through the paper. Known markings on the paper allow for image calibration. Noise reduction was attempted with removal of ‘inherent noise’more » from an open-field light image through the paper, but the method was found to be inconsequential. This is likely because the environment could not be controlled to the precision required for the sort of reproducible characterization of the quantum noise needed in order to meaningfully characterize and account for it. A multi-scale iterative edge detection algorithm was used for localizing the leaf ends. These were compared with the planned locations from the treatment console. Results: With a very basic setup, the image of the central bank A leaves 15–45, which are arguably the most important for beam modulation, differed from the planned location by [0.38±0.28] mm. Similarly, for bank B leaves 15–45 had a difference of [0.42±0.28] mm Conclusion: It should be possible to determine leaf position accurately with not much more than a modern hand-held camera and some software. This means we can have a periodic and independent verification of the dynalog file information. This is indicated by the precision already achieved using a basic setup and analysis methodology. Currently, work is being done to reduce imaging and setup errors, which will bring the leaf position error down further, and allow meaningful analysis over the full range of leaves.« less

  8. Conducted-Susceptibility Testing as an Alternative Approach to Unit-Level Radiated-Susceptibility Verifications

    NASA Astrophysics Data System (ADS)

    Badini, L.; Grassi, F.; Pignari, S. A.; Spadacini, G.; Bisognin, P.; Pelissou, P.; Marra, S.

    2016-05-01

    This work presents a theoretical rationale for the substitution of radiated-susceptibility (RS) verifications defined in current aerospace standards with an equivalent conducted-susceptibility (CS) test procedure based on bulk current injection (BCI) up to 500 MHz. Statistics is used to overcome the lack of knowledge about uncontrolled or uncertain setup parameters, with particular reference to the common-mode impedance of equipment. The BCI test level is properly investigated so to ensure correlation of currents injected in the equipment under test via CS and RS. In particular, an over-testing probability quantifies the severity of the BCI test with respect to the RS test.

  9. Technical note: patient-specific quality assurance methods for TomoDirect(TM) whole breast treatment delivery.

    PubMed

    Catuzzo, P; Zenone, F; Aimonetto, S; Peruzzo, A; Casanova Borca, V; Pasquino, M; Franco, P; La Porta, M R; Ricardi, U; Tofani, S

    2012-07-01

    To investigate the feasibility of implementing a novel approach for patient-specific QA of TomoDirect(TM) whole breast treatment. The most currently used TomoTherapy DQA method, consisting in the verification of the 2D dose distribution in a coronal or sagittal plane of the Cheese Phantom by means of gafchromic films, was compared with an alternative approach based on the use of two commercially available diode arrays, MapCHECK2(TM) and ArcCHECK(TM). The TomoDirect(TM) plans of twenty patients with a primary unilateral breast cancer were applied to a CT scan of the Cheese Phantom and a MVCT dataset of the diode arrays. Then measurements of 2D dose distribution were performed and compared with the calculated ones using the gamma analysis method with different sets of DTA and DD criteria (3%-3 mm, 3%-2 mm). The sensitivity of the diode arrays to detect delivery and setup errors was also investigated. The measured dose distributions showed excellent agreement with the TPS calculations for each detector, with averaged fractions of passed Γ values greater than 95%. The percentage of points satisfying the constraint Γ < 1 was significantly higher for MapCHECK2(TM) than for ArcCHECK(TM) and gafchromic films using both the 3%-3 mm and 3%-2 mm gamma criteria. Both the diode arrays show a good sensitivity to delivery and setup errors using a 3%-2 mm gamma criteria. MapCHECK2™ and ArcCHECK(TM) may fulfill the demands of an adequate system for TomoDirect(TM) patient-specific QA.

  10. Verification of respiratory-gated radiotherapy with new real-time tumour-tracking radiotherapy system using cine EPID images and a log file

    NASA Astrophysics Data System (ADS)

    Shiinoki, Takehiro; Hanazawa, Hideki; Yuasa, Yuki; Fujimoto, Koya; Uehara, Takuya; Shibuya, Keiko

    2017-02-01

    A combined system comprising the TrueBeam linear accelerator and a new real-time tumour-tracking radiotherapy system, SyncTraX, was installed at our institution. The objectives of this study are to develop a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine electronic portal image device (EPID) images and a log file and to verify this treatment in clinical cases. Respiratory-gated radiotherapy was performed using TrueBeam and the SyncTraX system. Cine EPID images and a log file were acquired for a phantom and three patients during the course of the treatment. Digitally reconstructed radiographs (DRRs) were created for each treatment beam using a planning CT set. The cine EPID images, log file, and DRRs were analysed using a developed software. For the phantom case, the accuracy of the proposed method was evaluated to verify the respiratory-gated radiotherapy. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker used as an internal surrogate were calculated to evaluate the gating accuracy and set-up uncertainty in the superior-inferior (SI), anterior-posterior (AP), and left-right (LR) directions. The proposed method achieved high accuracy for the phantom verification. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker were  ⩽3 mm and  ±3 mm in the SI, AP, and LR directions. We proposed a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine EPID images and a log file and showed that this treatment is performed with high accuracy in clinical cases. This work was partly presented at the 58th Annual meeting of American Association of Physicists in Medicine.

  11. Verification of respiratory-gated radiotherapy with new real-time tumour-tracking radiotherapy system using cine EPID images and a log file.

    PubMed

    Shiinoki, Takehiro; Hanazawa, Hideki; Yuasa, Yuki; Fujimoto, Koya; Uehara, Takuya; Shibuya, Keiko

    2017-02-21

    A combined system comprising the TrueBeam linear accelerator and a new real-time tumour-tracking radiotherapy system, SyncTraX, was installed at our institution. The objectives of this study are to develop a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine electronic portal image device (EPID) images and a log file and to verify this treatment in clinical cases. Respiratory-gated radiotherapy was performed using TrueBeam and the SyncTraX system. Cine EPID images and a log file were acquired for a phantom and three patients during the course of the treatment. Digitally reconstructed radiographs (DRRs) were created for each treatment beam using a planning CT set. The cine EPID images, log file, and DRRs were analysed using a developed software. For the phantom case, the accuracy of the proposed method was evaluated to verify the respiratory-gated radiotherapy. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker used as an internal surrogate were calculated to evaluate the gating accuracy and set-up uncertainty in the superior-inferior (SI), anterior-posterior (AP), and left-right (LR) directions. The proposed method achieved high accuracy for the phantom verification. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker were  ⩽3 mm and  ±3 mm in the SI, AP, and LR directions. We proposed a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine EPID images and a log file and showed that this treatment is performed with high accuracy in clinical cases.

  12. Improved human observer performance in digital reconstructed radiograph verification in head and neck cancer radiotherapy.

    PubMed

    Sturgeon, Jared D; Cox, John A; Mayo, Lauren L; Gunn, G Brandon; Zhang, Lifei; Balter, Peter A; Dong, Lei; Awan, Musaddiq; Kocak-Uzel, Esengul; Mohamed, Abdallah Sherif Radwan; Rosenthal, David I; Fuller, Clifton David

    2015-10-01

    Digitally reconstructed radiographs (DRRs) are routinely used as an a priori reference for setup correction in radiotherapy. The spatial resolution of DRRs may be improved to reduce setup error in fractionated radiotherapy treatment protocols. The influence of finer CT slice thickness reconstruction (STR) and resultant increased resolution DRRs on physician setup accuracy was prospectively evaluated. Four head and neck patient CT-simulation images were acquired and used to create DRR cohorts by varying STRs at 0.5, 1, 2, 2.5, and 3 mm. DRRs were displaced relative to a fixed isocenter using 0-5 mm random shifts in the three cardinal axes. Physician observers reviewed DRRs of varying STRs and displacements and then aligned reference and test DRRs replicating daily KV imaging workflow. A total of 1,064 images were reviewed by four blinded physicians. Observer errors were analyzed using nonparametric statistics (Friedman's test) to determine whether STR cohorts had detectably different displacement profiles. Post hoc bootstrap resampling was applied to evaluate potential generalizability. The observer-based trial revealed a statistically significant difference between cohort means for observer displacement vector error ([Formula: see text]) and for [Formula: see text]-axis [Formula: see text]. Bootstrap analysis suggests a 15% gain in isocenter translational setup error with reduction of STR from 3 mm to [Formula: see text]2 mm, though interobserver variance was a larger feature than STR-associated measurement variance. Higher resolution DRRs generated using finer CT scan STR resulted in improved observer performance at shift detection and could decrease operator-dependent geometric error. Ideally, CT STRs [Formula: see text]2 mm should be utilized for DRR generation in the head and neck.

  13. Clinical Implementation of a Model-Based In Vivo Dose Verification System for Stereotactic Body Radiation Therapy-Volumetric Modulated Arc Therapy Treatments Using the Electronic Portal Imaging Device.

    PubMed

    McCowan, Peter M; Asuni, Ganiyu; Van Uytven, Eric; VanBeek, Timothy; McCurdy, Boyd M C; Loewen, Shaun K; Ahmed, Naseer; Bashir, Bashir; Butler, James B; Chowdhury, Amitava; Dubey, Arbind; Leylek, Ahmet; Nashed, Maged

    2017-04-01

    To report findings from an in vivo dosimetry program implemented for all stereotactic body radiation therapy patients over a 31-month period and discuss the value and challenges of utilizing in vivo electronic portal imaging device (EPID) dosimetry clinically. From December 2013 to July 2016, 117 stereotactic body radiation therapy-volumetric modulated arc therapy patients (100 lung, 15 spine, and 2 liver) underwent 602 EPID-based in vivo dose verification events. A developed model-based dose reconstruction algorithm calculates the 3-dimensional dose distribution to the patient by back-projecting the primary fluence measured by the EPID during treatment. The EPID frame-averaging was optimized in June 2015. For each treatment, a 3%/3-mm γ comparison between our EPID-derived dose and the Eclipse AcurosXB-predicted dose to the planning target volume (PTV) and the ≥20% isodose volume were performed. Alert levels were defined as γ pass rates <85% (lung and liver) and <80% (spine). Investigations were carried out for all fractions exceeding the alert level and were classified as follows: EPID-related, algorithmic, patient setup, anatomic change, or unknown/unidentified errors. The percentages of fractions exceeding the alert levels were 22.6% for lung before frame-average optimization and 8.0% for lung, 20.0% for spine, and 10.0% for liver after frame-average optimization. Overall, mean (± standard deviation) planning target volume γ pass rates were 90.7% ± 9.2%, 87.0% ± 9.3%, and 91.2% ± 3.4% for the lung, spine, and liver patients, respectively. Results from the clinical implementation of our model-based in vivo dose verification method using on-treatment EPID images is reported. The method is demonstrated to be valuable for routine clinical use for verifying delivered dose as well as for detecting errors. Copyright © 2017 Elsevier Inc. All rights reserved.

  14. Clinical Implementation of a Model-Based In Vivo Dose Verification System for Stereotactic Body Radiation Therapy–Volumetric Modulated Arc Therapy Treatments Using the Electronic Portal Imaging Device

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    McCowan, Peter M., E-mail: pmccowan@cancercare.mb.ca; Asuni, Ganiyu; Van Uytven, Eric

    Purpose: To report findings from an in vivo dosimetry program implemented for all stereotactic body radiation therapy patients over a 31-month period and discuss the value and challenges of utilizing in vivo electronic portal imaging device (EPID) dosimetry clinically. Methods and Materials: From December 2013 to July 2016, 117 stereotactic body radiation therapy–volumetric modulated arc therapy patients (100 lung, 15 spine, and 2 liver) underwent 602 EPID-based in vivo dose verification events. A developed model-based dose reconstruction algorithm calculates the 3-dimensional dose distribution to the patient by back-projecting the primary fluence measured by the EPID during treatment. The EPID frame-averaging was optimized in Junemore » 2015. For each treatment, a 3%/3-mm γ comparison between our EPID-derived dose and the Eclipse AcurosXB–predicted dose to the planning target volume (PTV) and the ≥20% isodose volume were performed. Alert levels were defined as γ pass rates <85% (lung and liver) and <80% (spine). Investigations were carried out for all fractions exceeding the alert level and were classified as follows: EPID-related, algorithmic, patient setup, anatomic change, or unknown/unidentified errors. Results: The percentages of fractions exceeding the alert levels were 22.6% for lung before frame-average optimization and 8.0% for lung, 20.0% for spine, and 10.0% for liver after frame-average optimization. Overall, mean (± standard deviation) planning target volume γ pass rates were 90.7% ± 9.2%, 87.0% ± 9.3%, and 91.2% ± 3.4% for the lung, spine, and liver patients, respectively. Conclusions: Results from the clinical implementation of our model-based in vivo dose verification method using on-treatment EPID images is reported. The method is demonstrated to be valuable for routine clinical use for verifying delivered dose as well as for detecting errors.« less

  15. Monte Carlo simulations to replace film dosimetry in IMRT verification.

    PubMed

    Goetzfried, Thomas; Rickhey, Mark; Treutwein, Marius; Koelbl, Oliver; Bogner, Ludwig

    2011-01-01

    Patient-specific verification of intensity-modulated radiation therapy (IMRT) plans can be done by dosimetric measurements or by independent dose or monitor unit calculations. The aim of this study was the clinical evaluation of IMRT verification based on a fast Monte Carlo (MC) program with regard to possible benefits compared to commonly used film dosimetry. 25 head-and-neck IMRT plans were recalculated by a pencil beam based treatment planning system (TPS) using an appropriate quality assurance (QA) phantom. All plans were verified both by film and diode dosimetry and compared to MC simulations. The irradiated films, the results of diode measurements and the computed dose distributions were evaluated, and the data were compared on the basis of gamma maps and dose-difference histograms. Average deviations in the high-dose region between diode measurements and point dose calculations performed with the TPS and MC program were 0.7 ± 2.7% and 1.2 ± 3.1%, respectively. For film measurements, the mean gamma values with 3% dose difference and 3mm distance-to-agreement were 0.74 ± 0.28 (TPS as reference) with dose deviations up to 10%. Corresponding values were significantly reduced to 0.34 ± 0.09 for MC dose calculation. The total time needed for both verification procedures is comparable, however, by far less labor intensive in the case of MC simulations. The presented study showed that independent dose calculation verification of IMRT plans with a fast MC program has the potential to eclipse film dosimetry more and more in the near future. Thus, the linac-specific QA part will necessarily become more important. In combination with MC simulations and due to the simple set-up, point-dose measurements for dosimetric plausibility checks are recommended at least in the IMRT introduction phase. Copyright © 2010. Published by Elsevier GmbH.

  16. Sierra/SolidMechanics 4.48 Verification Tests Manual.

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Plews, Julia A.; Crane, Nathan K; de Frias, Gabriel Jose

    2018-03-01

    Presented in this document is a small portion of the tests that exist in the Sierra / SolidMechanics (Sierra / SM) verification test suite. Most of these tests are run nightly with the Sierra / SM code suite, and the results of the test are checked versus the correct analytical result. For each of the tests presented in this document, the test setup, a description of the analytic solution, and comparison of the Sierra / SM code results to the analytic solution is provided. Mesh convergence is also checked on a nightly basis for several of these tests. This documentmore » can be used to confirm that a given code capability is verified or referenced as a compilation of example problems. Additional example problems are provided in the Sierra / SM Example Problems Manual. Note, many other verification tests exist in the Sierra / SM test suite, but have not yet been included in this manual.« less

  17. Sierra/SolidMechanics 4.48 Verification Tests Manual.

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Plews, Julia A.; Crane, Nathan K.; de Frias, Gabriel Jose

    Presented in this document is a small portion of the tests that exist in the Sierra / SolidMechanics (Sierra / SM) verification test suite. Most of these tests are run nightly with the Sierra / SM code suite, and the results of the test are checked versus the correct analytical result. For each of the tests presented in this document, the test setup, a description of the analytic solution, and comparison of the Sierra / SM code results to the analytic solution is provided. Mesh convergence is also checked on a nightly basis for several of these tests. This documentmore » can be used to confirm that a given code capability is verified or referenced as a compilation of example problems. Additional example problems are provided in the Sierra / SM Example Problems Manual. Note, many other verification tests exist in the Sierra / SM test suite, but have not yet been included in this manual.« less

  18. Beam Loss Monitoring for LHC Machine Protection

    NASA Astrophysics Data System (ADS)

    Holzer, Eva Barbara; Dehning, Bernd; Effnger, Ewald; Emery, Jonathan; Grishin, Viatcheslav; Hajdu, Csaba; Jackson, Stephen; Kurfuerst, Christoph; Marsili, Aurelien; Misiowiec, Marek; Nagel, Markus; Busto, Eduardo Nebot Del; Nordt, Annika; Roderick, Chris; Sapinski, Mariusz; Zamantzas, Christos

    The energy stored in the nominal LHC beams is two times 362 MJ, 100 times the energy of the Tevatron. As little as 1 mJ/cm3 deposited energy quenches a magnet at 7 TeV and 1 J/cm3 causes magnet damage. The beam dumps are the only places to safely dispose of this beam. One of the key systems for machine protection is the beam loss monitoring (BLM) system. About 3600 ionization chambers are installed at likely or critical loss locations around the LHC ring. The losses are integrated in 12 time intervals ranging from 40 μs to 84 s and compared to threshold values defined in 32 energy ranges. A beam abort is requested when potentially dangerous losses are detected or when any of the numerous internal system validation tests fails. In addition, loss data are used for machine set-up and operational verifications. The collimation system for example uses the loss data for set-up and regular performance verification. Commissioning and operational experience of the BLM are presented: The machine protection functionality of the BLM system has been fully reliable; the LHC availability has not been compromised by false beam aborts.

  19. SU-E-J-119: What Effect Have the Volume Defined in the Alignment Clipbox for Cervical Cancer Using Automatic Registration Methods for Cone- Beam CT Verification?

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wang, W; Yang, H; Wang, Y

    2014-06-01

    Purpose: To investigate the impact of different clipbox volumes with automated registration techniques using commercially available software with on board volumetric imaging(OBI) for treatment verification in cervical cancer patients. Methods: Fifty cervical cancer patients received daily CBCT scans(on-board imaging v1.5 system, Varian Medical Systems) during the first treatment week and weekly thereafter were included this analysis. A total of 450 CBCT scans were registered to the planning CTscan using pelvic clipbox(clipbox-Pelvic) and around PTV clip box(clipbox- PTV). The translations(anterior-posterior, left-right, superior-inferior) and the rotations(yaw, pitch and roll) errors for each matches were recorded. The setup errors and the systematic andmore » random errors for both of the clip-boxes were calculated. Paired Samples t test was used to analysis the differences between clipbox-Pelvic and clipbox-PTV. Results: . The SD of systematic error(σ) was 1.0mm, 2.0mm,3.2mm and 1.9mm,2.3mm, 3.0mm in the AP, LR and SI directions for clipbox-Pelvic and clipbox-PTV, respectively. The average random error(Σ)was 1.7mm, 2.0mm,4.2mm and 1.7mm,3.4mm, 4.4mm in the AP, LR and SI directions for clipbox-Pelvic and clipbox-PTV, respectively. But, only the SI direction was acquired significantly differences between two image registration volumes(p=0.002,p=0.01 for mean and SD). For rotations, the yaw mean/SD and the pitch SD were acquired significantly differences between clipbox-Pelvic and clipbox-PTV. Conclusion: The defined volume for Image registration is important for cervical cancer when 3D/3D match was used. The alignment clipbox can effect the setup errors obtained. Further analysis is need to determine the optimal defined volume to use the image registration in cervical cancer. Conflict of interest: none.« less

  20. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lin, T; Eldib, A; Hossain, M

    Purpose: Patient in-vivo measurements report lower readings than those predicted from TMR-based treatment planning on TBI patient knees and ankles where rice was placed to fill the gap between patient’s legs. This study is to understand and correct the under dosage of Total Body Irradiation(TBI) with rice tissue equivalent bolus placement at TBI treatment patient setup. Methods: Bilateral TBI scheme was investigated with rice bags bolus placing between patient’s two legs acting as missing tissue. In-house TMR based treatment planning system was commissioned with measurements under TBI condition at 10MV, i.e. source-to-reference distance 383.4cm with 40×40cm field size with 1cmmore » thickness Lucite. Predictions of patient specific dose points are reported at different sites with 200cGy prescription at patient umbilicus point. Solid water and rice bag phantoms are used at TBI conditions for the attenuation factor verification and CT scanned to verify the CT number and electron density. Results: We found that the rice bag bolus overall density is 11% lower than the water; however, the attenuation factor of rice bags could become 15% lower than that of water at TBI condition. This overestimate of rice bag electron density could cause the lack of lateral scatter and the lack of backscatter. This could Result in an overestimate of dose at in-vivo dosimeter measurement points with TMR-based treatment planning systems. Observations of patient specific optically stimulated luminescent dosimeters(OSLDs) were used to confirm this overestimation. Measurements of setups with increasing the rice bag filled patient leg separation were performed to demonstrate eliminating the overdose issue. Conclusion: Rice bolus has a lower electron density than water does(11%) but results in 15% lower in attenuation factor at TBI condition. This effect was observed in patient delivery with OSLD measurements and can be corrected by increasing the filling rice bolus thickness with 15% longer of separation.« less

  1. Automatic x-ray image contrast enhancement based on parameter auto-optimization.

    PubMed

    Qiu, Jianfeng; Harold Li, H; Zhang, Tiezhi; Ma, Fangfang; Yang, Deshan

    2017-11-01

    Insufficient image contrast associated with radiation therapy daily setup x-ray images could negatively affect accurate patient treatment setup. We developed a method to perform automatic and user-independent contrast enhancement on 2D kilo voltage (kV) and megavoltage (MV) x-ray images. The goal was to provide tissue contrast optimized for each treatment site in order to support accurate patient daily treatment setup and the subsequent offline review. The proposed method processes the 2D x-ray images with an optimized image processing filter chain, which consists of a noise reduction filter and a high-pass filter followed by a contrast limited adaptive histogram equalization (CLAHE) filter. The most important innovation is to optimize the image processing parameters automatically to determine the required image contrast settings per disease site and imaging modality. Three major parameters controlling the image processing chain, i.e., the Gaussian smoothing weighting factor for the high-pass filter, the block size, and the clip limiting parameter for the CLAHE filter, were determined automatically using an interior-point constrained optimization algorithm. Fifty-two kV and MV x-ray images were included in this study. The results were manually evaluated and ranked with scores from 1 (worst, unacceptable) to 5 (significantly better than adequate and visually praise worthy) by physicians and physicists. The average scores for the images processed by the proposed method, the CLAHE, and the best window-level adjustment were 3.92, 2.83, and 2.27, respectively. The percentage of the processed images received a score of 5 were 48, 29, and 18%, respectively. The proposed method is able to outperform the standard image contrast adjustment procedures that are currently used in the commercial clinical systems. When the proposed method is implemented in the clinical systems as an automatic image processing filter, it could be useful for allowing quicker and potentially more accurate treatment setup and facilitating the subsequent offline review and verification. © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  2. A technique for pediatric total skin electron irradiation.

    PubMed

    Bao, Qinan; Hrycushko, Brian A; Dugas, Joseph P; Hager, Frederick H; Solberg, Timothy D

    2012-03-20

    Total skin electron irradiation (TSEI) is a special radiotherapy technique which has generally been used for treating adult patients with mycosis fungoides. Recently, two infants presented with leukemia cutis isolated to the skin requiring TSEI. This work discusses the commissioning and quality assurance (QA) methods for implementing a modified Stanford technique using a rotating harness system to position sedated pediatric patients treated with electrons to the total skin. Commissioning of pediatric TSEI consisted of absolute calibration, measurement of dosimetric parameters, and subsequent verification in a pediatric patient sized cylindrical phantom using radiographic film and optically stimulated luminance (OSL) dosimeters. The depth of dose penetration under TSEI treatment condition was evaluated using radiographic film sandwiched in the phantom and demonstrated a 2 cm penetration depth with the maximum dose located at the phantom surface. Dosimetry measurements on the cylindrical phantom and in-vivo measurements from the patients suggested that, the factor relating the skin and calibration point doses (i.e., the B-factor) was larger for the pediatric TSEI treatments as compared to adult TSEI treatments. Custom made equipment, including a rotating plate and harness, was fabricated and added to a standard total body irradiation stand and tested to facilitate patient setup under sedated condition. A pediatric TSEI QA program, consisting of daily output, energy, flatness, and symmetry measurements as well as in-vivo dosimetry verification for the first cycle was developed. With a long interval between pediatric TSEI cases, absolute dosimetry was also repeated as part of the QA program. In-vivo dosimetry for the first two infants showed that a dose of ± 10% of the prescription dose can be achieved over the entire patient body. Though pediatric leukemia cutis and the subsequent need for TSEI are rare, the ability to commission the technique on a modified TBI stand is appealing for clinical implementation and has been successfully used for the treatment of two pediatric patients at our institution.

  3. A technique for pediatric total skin electron irradiation

    PubMed Central

    2012-01-01

    Background Total skin electron irradiation (TSEI) is a special radiotherapy technique which has generally been used for treating adult patients with mycosis fungoides. Recently, two infants presented with leukemia cutis isolated to the skin requiring TSEI. This work discusses the commissioning and quality assurance (QA) methods for implementing a modified Stanford technique using a rotating harness system to position sedated pediatric patients treated with electrons to the total skin. Methods and Results Commissioning of pediatric TSEI consisted of absolute calibration, measurement of dosimetric parameters, and subsequent verification in a pediatric patient sized cylindrical phantom using radiographic film and optically stimulated luminance (OSL) dosimeters. The depth of dose penetration under TSEI treatment condition was evaluated using radiographic film sandwiched in the phantom and demonstrated a 2 cm penetration depth with the maximum dose located at the phantom surface. Dosimetry measurements on the cylindrical phantom and in-vivo measurements from the patients suggested that, the factor relating the skin and calibration point doses (i.e., the B-factor) was larger for the pediatric TSEI treatments as compared to adult TSEI treatments. Custom made equipment, including a rotating plate and harness, was fabricated and added to a standard total body irradiation stand and tested to facilitate patient setup under sedated condition. A pediatric TSEI QA program, consisting of daily output, energy, flatness, and symmetry measurements as well as in-vivo dosimetry verification for the first cycle was developed. With a long interval between pediatric TSEI cases, absolute dosimetry was also repeated as part of the QA program. In-vivo dosimetry for the first two infants showed that a dose of ± 10% of the prescription dose can be achieved over the entire patient body. Conclusion Though pediatric leukemia cutis and the subsequent need for TSEI are rare, the ability to commission the technique on a modified TBI stand is appealing for clinical implementation and has been successfully used for the treatment of two pediatric patients at our institution. PMID:22433063

  4. The future of image-guided radiotherapy will be MR guided

    PubMed Central

    Wen, Zhifei; Sadagopan, Ramaswamy; Wang, Jihong; Ibbott, Geoffrey S

    2017-01-01

    Advances in image-guided radiotherapy (RT) have allowed for dose escalation and more precise radiation treatment delivery. Each decade brings new imaging technologies to help improve RT patient setup. Currently, the most frequently used method of three-dimensional pre-treatment image verification is performed with cone beam CT. However, more recent developments have provided RT with the ability to have on-board MRI coupled to the teleradiotherapy unit. This latest tool for treating cancer is known as MR-guided RT. Several varieties of these units have been designed and installed in centres across the globe. Their prevalence, history, advantages and disadvantages are discussed in this review article. In preparation for the next generation of image-guided RT, this review also covers where MR-guided RT might be heading in the near future. PMID:28256898

  5. Direct and full-scale experimental verifications towards ground-satellite quantum key distribution

    NASA Astrophysics Data System (ADS)

    Wang, Jian-Yu; Yang, Bin; Liao, Sheng-Kai; Zhang, Liang; Shen, Qi; Hu, Xiao-Fang; Wu, Jin-Cai; Yang, Shi-Ji; Jiang, Hao; Tang, Yan-Lin; Zhong, Bo; Liang, Hao; Liu, Wei-Yue; Hu, Yi-Hua; Huang, Yong-Mei; Qi, Bo; Ren, Ji-Gang; Pan, Ge-Sheng; Yin, Juan; Jia, Jian-Jun; Chen, Yu-Ao; Chen, Kai; Peng, Cheng-Zhi; Pan, Jian-Wei

    2013-05-01

    Quantum key distribution (QKD) provides the only intrinsically unconditional secure method for communication based on the principle of quantum mechanics. Compared with fibre-based demonstrations, free-space links could provide the most appealing solution for communication over much larger distances. Despite significant efforts, all realizations to date rely on stationary sites. Experimental verifications are therefore extremely crucial for applications to a typical low Earth orbit satellite. To achieve direct and full-scale verifications of our set-up, we have carried out three independent experiments with a decoy-state QKD system, and overcome all conditions. The system is operated on a moving platform (using a turntable), on a floating platform (using a hot-air balloon), and with a high-loss channel to demonstrate performances under conditions of rapid motion, attitude change, vibration, random movement of satellites, and a high-loss regime. The experiments address wide ranges of all leading parameters relevant to low Earth orbit satellites. Our results pave the way towards ground-satellite QKD and a global quantum communication network.

  6. Automatic Emboli Detection System for the Artificial Heart

    NASA Astrophysics Data System (ADS)

    Steifer, T.; Lewandowski, M.; Karwat, P.; Gawlikowski, M.

    In spite of the progress in material engineering and ventricular assist devices construction, thromboembolism remains the most crucial problem in mechanical heart supporting systems. Therefore, the ability to monitor the patient's blood for clot formation should be considered an important factor in development of heart supporting systems. The well-known methods for automatic embolus detection are based on the monitoring of the ultrasound Doppler signal. A working system utilizing ultrasound Doppler is being developed for the purpose of flow estimation and emboli detection in the clinical artificial heart ReligaHeart EXT. Thesystem will be based on the existing dual channel multi-gate Doppler device with RF digital processing. A specially developed clamp-on cannula probe, equipped with 2 - 4 MHz piezoceramic transducers, enables easy system setup. We present the issuesrelated to the development of automatic emboli detection via Doppler measurements. We consider several algorithms for the flow estimation and emboli detection. We discuss their efficiency and confront them with the requirements of our experimental setup. Theoretical considerations are then met with preliminary experimental findings from a) flow studies with blood mimicking fluid and b) in-vitro flow studies with animal blood. Finally, we discuss some more methodological issues - we consider several possible approaches to the problem of verification of the accuracy of the detection system.

  7. Verification of the Sentinel-4 focal plane subsystem

    NASA Astrophysics Data System (ADS)

    Williges, Christian; Uhlig, Mathias; Hilbert, Stefan; Rossmann, Hannes; Buchwinkler, Kevin; Babben, Steffen; Sebastian, Ilse; Hohn, Rüdiger; Reulke, Ralf

    2017-09-01

    The Sentinel-4 payload is a multi-spectral camera system, designed to monitor atmospheric conditions over Europe from a geostationary orbit. The German Aerospace Center, DLR Berlin, conducted the verification campaign of the Focal Plane Subsystem (FPS) during the second half of 2016. The FPS consists, of two Focal Plane Assemblies (FPAs), two Front End Electronics (FEEs), one Front End Support Electronic (FSE) and one Instrument Control Unit (ICU). The FPAs are designed for two spectral ranges: UV-VIS (305 nm - 500 nm) and NIR (750 nm - 775 nm). In this publication, we will present in detail the set-up of the verification campaign of the Sentinel-4 Qualification Model (QM). This set up will also be used for the upcoming Flight Model (FM) verification, planned for early 2018. The FPAs have to be operated at 215 K +/- 5 K, making it necessary to exploit a thermal vacuum chamber (TVC) for the test accomplishment. The test campaign consists mainly of radiometric tests. This publication focuses on the challenge to remotely illuminate both Sentinel-4 detectors as well as a reference detector homogeneously over a distance of approximately 1 m from outside the TVC. Selected test analyses and results will be presented.

  8. Robust Proton Pencil Beam Scanning Treatment Planning for Rectal Cancer Radiation Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Blanco Kiely, Janid Patricia, E-mail: jkiely@sas.upenn.edu; White, Benjamin M.

    2016-05-01

    Purpose: To investigate, in a treatment plan design and robustness study, whether proton pencil beam scanning (PBS) has the potential to offer advantages, relative to interfraction uncertainties, over photon volumetric modulated arc therapy (VMAT) in a locally advanced rectal cancer patient population. Methods and Materials: Ten patients received a planning CT scan, followed by an average of 4 weekly offline CT verification CT scans, which were rigidly co-registered to the planning CT. Clinical PBS plans were generated on the planning CT, using a single-field uniform-dose technique with single-posterior and parallel-opposed (LAT) fields geometries. The VMAT plans were generated on the planningmore » CT using 2 6-MV, 220° coplanar arcs. Clinical plans were forward-calculated on verification CTs to assess robustness relative to anatomic changes. Setup errors were assessed by forward-calculating clinical plans with a ±5-mm (left–right, anterior–posterior, superior–inferior) isocenter shift on the planning CT. Differences in clinical target volume and organ at risk dose–volume histogram (DHV) indicators between plans were tested for significance using an appropriate Wilcoxon test (P<.05). Results: Dosimetrically, PBS plans were statistically different from VMAT plans, showing greater organ at risk sparing. However, the bladder was statistically identical among LAT and VMAT plans. The clinical target volume coverage was statistically identical among all plans. The robustness test found that all DVH indicators for PBS and VMAT plans were robust, except the LAT's genitalia (V5, V35). The verification CT plans showed that all DVH indicators were robust. Conclusions: Pencil beam scanning plans were found to be as robust as VMAT plans relative to interfractional changes during treatment when posterior beam angles and appropriate range margins are used. Pencil beam scanning dosimetric gains in the bowel (V15, V20) over VMAT suggest that using PBS to treat rectal cancer may reduce radiation treatment–related toxicity.« less

  9. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Saleh, H; Ferjani, S; Masssey, V

    Purpose: Perform dosimetric comparison between planned and delivered dose in the junction area, measure daily dose variation in the arc junction area for pediatric patients treated for medulloblastoma using Craniospinal axis irradiation(CSI) Material and methods Dose comparison in the junction area, daily dose variation in the arc junction area for a Rando Phantom and 5 pediatric patients treated using CSI technique were analyzed. Plans were created using the Eclipse treatment planning system. Two arcs for cranium and 1 arc for spine region were used. Planar dose matrix was created by projecting phantom and patient plan into the ArcCheck phantom. EBT3more » film was placed in the middle of ArcCheck plug to measure dose distribution in the junction areaDuring patient treatment, strip of EBT3 film was placed daily at each junction area for verification. EBT3 films were scanned using a flatbed scanner, Epson Expression 10000 XL. Film QA pro software was used to analyze film. Scanning and analysis was performed according to vendor recommendations and AAPM TG-55 report. Films were scanned and analyzed daily after each treatment and at the end of treatment course. Planar dose distributions from films were compared with planar dose distribution from treatment planning system. Results: Comparison of planned vs. measured dose distributions for patients have passing rates of 90%–100% with 3% and 3 mm gamma analysis. In some of the treatment fractions, daily setup film showed variation in dose distribution in the junction area. Conclusion: It is critical to measure dose distribution in the arc junction area and use additional quality assurance measures to verify daily setup for CSI patient where one or more junctions are present. EBT3 film prove to be a good tool to achieve this task considering flexibility associated with the film such as symmetry, self-developing and ease of use.« less

  10. A Mock Circulatory System Incorporating a Compliant 3D-Printed Anatomical Model to Investigate Pulmonary Hemodynamics.

    PubMed

    Knoops, Paul G M; Biglino, Giovanni; Hughes, Alun D; Parker, Kim H; Xu, Linzhang; Schievano, Silvia; Torii, Ryo

    2017-07-01

    A realistic mock circulatory system (MCS) could be a valuable in vitro testbed to study human circulatory hemodynamics. The objective of this study was to design a MCS replicating the pulmonary arterial circulation, incorporating an anatomically representative arterial model suitable for testing clinically relevant scenarios. A second objective of the study was to ensure the system's compatibility with magnetic resonance imaging (MRI) for additional measurements. A latex pulmonary arterial model with two generations of bifurcations was manufactured starting from a 3D-printed mold reconstructed from patient data. The model was incorporated into a MCS for in vitro hydrodynamic measurements. The setup was tested under physiological pulsatile flow conditions and results were evaluated using wave intensity analysis (WIA) to investigate waves traveling in the arterial system. Increased pulmonary vascular resistance (IPVR) was simulated as an example of one pathological scenario. Flow split between right and left pulmonary artery was found to be realistic (54 and 46%, respectively). No substantial difference in pressure waveform was observed throughout the various generations of bifurcations. Based on WIA, three main waves were identified in the main pulmonary artery (MPA), that is, forward compression wave, backward compression wave, and forward expansion wave. For IPVR, a rise in mean pressure was recorded in the MPA, within the clinical range of pulmonary arterial hypertension. The feasibility of using the MCS in the MRI scanner was demonstrated with the MCS running 2 h consecutively while acquiring preliminary MRI data. This study shows the development and verification of a pulmonary MCS, including an anatomically correct, compliant latex phantom. The setup can be useful to explore a wide range of hemodynamic questions, including the development of patient- and pathology-specific models, considering the ease and low cost of producing rapid prototyping molds, and the versatility of the setup for invasive and noninvasive (i.e., MRI) measurements. © 2016 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

  11. A Mock Circulatory System Incorporating a Compliant 3D-Printed Anatomical Model to Investigate Pulmonary Hemodynamics

    PubMed Central

    Knoops, Paul G.M.; Biglino, Giovanni; Hughes, Alun D.; Parker, Kim H.; Xu, Linzhang; Schievano, Silvia; Torii, Ryo

    2017-01-01

    A realistic mock circulatory system (MCS) could be a valuable in vitro testbed to study human circulatory hemodynamics. The objective of this study was to design a MCS replicating the pulmonary arterial circulation, incorporating an anatomically representative arterial model suitable for testing clinically relevant scenarios. A second objective of the study was to ensure the system's compatibility with magnetic resonance imaging (MRI) for additional measurements. A latex pulmonary arterial model with two generations of bifurcations was manufactured starting from a 3D-printed mold reconstructed from patient data. The model was incorporated into a MCS for in vitro hydrodynamic measurements. The setup was tested under physiological pulsatile flow conditions and results were evaluated using wave intensity analysis (WIA) to investigate waves traveling in the arterial system. Increased pulmonary vascular resistance (IPVR) was simulated as an example of one pathological scenario. Flow split between right and left pulmonary artery was found to be realistic (54 and 46%, respectively). No substantial difference in pressure waveform was observed throughout the various generations of bifurcations. Based on WIA, three main waves were identified in the main pulmonary artery (MPA), that is, forward compression wave, backward compression wave, and forward expansion wave. For IPVR, a rise in mean pressure was recorded in the MPA, within the clinical range of pulmonary arterial hypertension. The feasibility of using the MCS in the MRI scanner was demonstrated with the MCS running 2 h consecutively while acquiring preliminary MRI data. This study shows the development and verification of a pulmonary MCS, including an anatomically correct, compliant latex phantom. The setup can be useful to explore a wide range of hemodynamic questions, including the development of patient- and pathology-specific models, considering the ease and low cost of producing rapid prototyping molds, and the versatility of the setup for invasive and noninvasive (i.e., MRI) measurements. PMID:27925228

  12. A high sensitivity heterodyne interferometer as a possible optical readout for the LISA gravitational reference sensor and its application to technology verification

    NASA Astrophysics Data System (ADS)

    Gohlke, Martin; Schuldt, Thilo; Weise, Dennis; Cordero, Jorge; Peters, Achim; Johann, Ulrich; Braxmaier, Claus

    2017-11-01

    The gravitational wave detector LISA utilizes as current baseline a high sensitivity Optical Readout (ORO) for measuring the relative position and tilt of a free flying proof mass with respect to the satellite housing. The required sensitivities in the frequency band from 30 μHz to 1Hz are ˜ pm/ √ Hz for the translation√ and nrad/√ Hz for the tilt measurement. EADS Astrium, in collaboration with the Humboldt University Berlin and the University of Applied Sciences Konstanz, has realized a prototype ORO over the past years. The interferometer is based on a highly symmetric design where both, measurement and reference beam have a similar optical pathlength, and the same frequency and polarization. The technique of differential wavefront sensing (DWS) for tilt measurement is implemented. With our setup noise levels below 5pm/ √Hz for translation and below 10nrad/ √Hz for tilt measurements - both for frequencies above 10mHz - were demonstrated. We give an overview over the experimental setup, its current performance and the planned improvements. We also discuss the application to first verification of critical LISA aspects. As example we present measurements of the coefficient of thermal expansion (CTE) of various carbon fiber reinforced plastic (CFRP) including a "near-zero-CTE" tube.

  13. Communication Architecture in Mixed-Reality Simulations of Unmanned Systems

    PubMed Central

    2018-01-01

    Verification of the correct functionality of multi-vehicle systems in high-fidelity scenarios is required before any deployment of such a complex system, e.g., in missions of remote sensing or in mobile sensor networks. Mixed-reality simulations where both virtual and physical entities can coexist and interact have been shown to be beneficial for development, testing, and verification of such systems. This paper deals with the problems of designing a certain communication subsystem for such highly desirable realistic simulations. Requirements of this communication subsystem, including proper addressing, transparent routing, visibility modeling, or message management, are specified prior to designing an appropriate solution. Then, a suitable architecture of this communication subsystem is proposed together with solutions to the challenges that arise when simultaneous virtual and physical message transmissions occur. The proposed architecture can be utilized as a high-fidelity network simulator for vehicular systems with implicit mobility models that are given by real trajectories of the vehicles. The architecture has been utilized within multiple projects dealing with the development and practical deployment of multi-UAV systems, which support the architecture’s viability and advantages. The provided experimental results show the achieved similarity of the communication characteristics of the fully deployed hardware setup to the setup utilizing the proposed mixed-reality architecture. PMID:29538290

  14. Communication Architecture in Mixed-Reality Simulations of Unmanned Systems.

    PubMed

    Selecký, Martin; Faigl, Jan; Rollo, Milan

    2018-03-14

    Verification of the correct functionality of multi-vehicle systems in high-fidelity scenarios is required before any deployment of such a complex system, e.g., in missions of remote sensing or in mobile sensor networks. Mixed-reality simulations where both virtual and physical entities can coexist and interact have been shown to be beneficial for development, testing, and verification of such systems. This paper deals with the problems of designing a certain communication subsystem for such highly desirable realistic simulations. Requirements of this communication subsystem, including proper addressing, transparent routing, visibility modeling, or message management, are specified prior to designing an appropriate solution. Then, a suitable architecture of this communication subsystem is proposed together with solutions to the challenges that arise when simultaneous virtual and physical message transmissions occur. The proposed architecture can be utilized as a high-fidelity network simulator for vehicular systems with implicit mobility models that are given by real trajectories of the vehicles. The architecture has been utilized within multiple projects dealing with the development and practical deployment of multi-UAV systems, which support the architecture's viability and advantages. The provided experimental results show the achieved similarity of the communication characteristics of the fully deployed hardware setup to the setup utilizing the proposed mixed-reality architecture.

  15. Fingerprint changes and verification failure among patients with hand dermatitis.

    PubMed

    Lee, Chew Kek; Chang, Choong Chor; Johar, Asmah; Puwira, Othman; Roshidah, Baba

    2013-03-01

    To determine the prevalence of fingerprint verification failure and to define and quantify the fingerprint changes associated with fingerprint verification failure. Case-control study. Referral public dermatology center. The study included 100 consecutive patients with clinical hand dermatitis involving the palmar distal phalanx of either thumb and 100 age-, sex-, and ethnicity-matched controls. Patients with an altered thumb print due to other causes and palmar hyperhidrosis were excluded. Fingerprint verification(pass/fail) and hand eczema severity index score. Twenty-seven percent of patients failed fingerprint verification compared with 2% of controls. Fingerprint verification failure was associated with a higher hand eczema severity index score (P.001). The main fingerprint abnormalities were fingerprint dystrophy (42.0%) and abnormal white lines (79.5%). The number of abnormal white lines was significantly higher among the patients with hand dermatitis compared with controls(P=.001). Among the patients with hand dermatitis, theodds of failing fingerprint verification with fingerprint dystrophy was 4.01. The presence of broad lines and long lines was associated with a greater odds of fingerprint verification failure (odds ratio [OR], 8.04; 95% CI, 3.56-18.17 and OR, 2.37; 95% CI, 1.31-4.27, respectively),while the presence of thin lines was protective of verification failure (OR, 0.45; 95% CI, 0.23-0.89). Fingerprint verification failure is a significant problem among patients with more severe hand dermatitis. It is mainly due to fingerprint dystrophy and abnormal white lines. Malaysian National Medical Research Register Identifier: NMRR-11-30-8226

  16. Evaluation of overall setup accuracy and adequate setup margins in pelvic image-guided radiotherapy: Comparison of the male and female patients

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Laaksomaa, Marko, E-mail: marko.laaksomaa@pshp.fi; Kapanen, Mika; Department of Medical Physics, Tampere University Hospital

    We evaluated adequate setup margins for the radiotherapy (RT) of pelvic tumors based on overall position errors of bony landmarks. We also estimated the difference in setup accuracy between the male and female patients. Finally, we compared the patient rotation for 2 immobilization devices. The study cohort included consecutive 64 male and 64 female patients. Altogether, 1794 orthogonal setup images were analyzed. Observer-related deviation in image matching and the effect of patient rotation were explicitly determined. Overall systematic and random errors were calculated in 3 orthogonal directions. Anisotropic setup margins were evaluated based on residual errors after weekly image guidance.more » The van Herk formula was used to calculate the margins. Overall, 100 patients were immobilized with a house-made device. The patient rotation was compared against 28 patients immobilized with CIVCO's Kneefix and Feetfix. We found that the usually applied isotropic setup margin of 8 mm covered all the uncertainties related to patient setup for most RT treatments of the pelvis. However, margins of even 10.3 mm were needed for the female patients with very large pelvic target volumes centered either in the symphysis or in the sacrum containing both of these structures. This was because the effect of rotation (p ≤ 0.02) and the observer variation in image matching (p ≤ 0.04) were significantly larger for the female patients than for the male patients. Even with daily image guidance, the required margins remained larger for the women. Patient rotations were largest about the lateral axes. The difference between the required margins was only 1 mm for the 2 immobilization devices. The largest component of overall systematic position error came from patient rotation. This emphasizes the need for rotation correction. Overall, larger position errors and setup margins were observed for the female patients with pelvic cancer than for the male patients.« less

  17. Range verification for eye proton therapy based on proton-induced x-ray emissions from implanted metal markers

    NASA Astrophysics Data System (ADS)

    La Rosa, Vanessa; Kacperek, Andrzej; Royle, Gary; Gibson, Adam

    2014-06-01

    Metal fiducial markers are often implanted on the back of the eye before proton therapy to improve target localization and reduce patient setup errors. We aim to detect characteristic x-ray emissions from metal targets during proton therapy to verify the treatment range accuracy. Initially gold was chosen for its biocompatibility properties. Proton-induced x-ray emissions (PIXE) from a 15 mm diameter gold marker were detected at different penetration depths of a 59 MeV proton beam at the CATANA proton facility at INFN-LNS (Italy). The Monte Carlo code Geant4 was used to reproduce the experiment and to investigate the effect of different size markers, materials, and the response to both mono-energetic and fully modulated beams. The intensity of the emitted x-rays decreases with decreasing proton energy and thus decreases with depth. If we assume the range to be the depth at which the dose is reduced to 10% of its maximum value and we define the residual range as the distance between the marker and the range of the beam, then the minimum residual range which can be detected with 95% confidence level is the depth at which the PIXE peak is equal to 1.96 σbkg, which is the standard variation of the background noise. With our system and experimental setup this value is 3 mm, when 20 GyE are delivered to a gold marker of 15 mm diameter. Results from silver are more promising. Even when a 5 mm diameter silver marker is placed at a depth equal to the range, the PIXE peak is 2.1 σbkg. Although these quantitative results are dependent on the experimental setup used in this research study, they demonstrate that the real-time analysis of the PIXE emitted by fiducial metal markers can be used to derive beam range. Further analysis are needed to demonstrate the feasibility of the technique in a clinical setup.

  18. Data capture by digital pen in clinical trials: a qualitative and quantitative study.

    PubMed

    Estellat, Candice; Tubach, Florence; Costa, Yolande; Hoffmann, Isabelle; Mantz, Jean; Ravaud, Philippe

    2008-05-01

    To investigate the use of the digital pen (DP) system to collect data in a clinical trial. To assess the accuracy of the system in this setting. Qualitative study based on semistructured interviews and a focus group. Quantitative study comparing the DP system and a double manual data-entry system in accuracy of acquiring data by variable type (tick boxes, dates, numbers, letters). An ongoing randomised multicentric clinical trial in tertiary care in France. 27 investigators involved in the trial (anaesthetists) who did or did not include patients, 4 study monitors and the study coordinator. Six key findings emerged: 1) the DP system was easy to use; its utilisation was intuitive, even for investigators inexperienced in informatics; 2) despite its portability, the DP was not always used in front of patients; 3) the DP system did not affect patient recruitment; 4) most of the technical problems of the system occurred during setup (compatibility, password access, antivirus software); 5) the main advantage was quickness of data availability for the study coordination staff and the main hindrance was the extra time required for online verification; and 6) all investigators were ready to use the system again. The investigators had to check 16% of data obtained by the DP system during the verification step. There is no relevant difference between the number of errors for the DP and the double manual data-entry systems: 8/5022 versus 6/5022 data entries. 5 out of 8 DP-system failures were due to the intelligent character recognition system. The DP system has a good acceptability among all investigators in a clinical setting, whether they are experienced with computers or not, and a good accuracy, as compared with double manual data entry.

  19. Efficacy and workload analysis of a fixed vertical couch position technique and a fixed‐action–level protocol in whole‐breast radiotherapy

    PubMed Central

    Verhoeven, Karolien; Weltens, Caroline; Van den Heuvel, Frank

    2015-01-01

    Quantification of the setup errors is vital to define appropriate setup margins preventing geographical misses. The no‐action–level (NAL) correction protocol reduces the systematic setup errors and, hence, the setup margins. The manual entry of the setup corrections in the record‐and‐verify software, however, increases the susceptibility of the NAL protocol to human errors. Moreover, the impact of the skin mobility on the anteroposterior patient setup reproducibility in whole‐breast radiotherapy (WBRT) is unknown. In this study, we therefore investigated the potential of fixed vertical couch position‐based patient setup in WBRT. The possibility to introduce a threshold for correction of the systematic setup errors was also explored. We measured the anteroposterior, mediolateral, and superior–inferior setup errors during fractions 1–12 and weekly thereafter with tangential angled single modality paired imaging. These setup data were used to simulate the residual setup errors of the NAL protocol, the fixed vertical couch position protocol, and the fixed‐action–level protocol with different correction thresholds. Population statistics of the setup errors of 20 breast cancer patients and 20 breast cancer patients with additional regional lymph node (LN) irradiation were calculated to determine the setup margins of each off‐line correction protocol. Our data showed the potential of the fixed vertical couch position protocol to restrict the systematic and random anteroposterior residual setup errors to 1.8 mm and 2.2 mm, respectively. Compared to the NAL protocol, a correction threshold of 2.5 mm reduced the frequency of mediolateral and superior–inferior setup corrections with 40% and 63%, respectively. The implementation of the correction threshold did not deteriorate the accuracy of the off‐line setup correction compared to the NAL protocol. The combination of the fixed vertical couch position protocol, for correction of the anteroposterior setup error, and the fixed‐action–level protocol with 2.5 mm correction threshold, for correction of the mediolateral and the superior–inferior setup errors, was proved to provide adequate and comparable patient setup accuracy in WBRT and WBRT with additional LN irradiation. PACS numbers: 87.53.Kn, 87.57.‐s

  20. SU-E-J-15: A Patient-Centered Scheme to Mitigate Impacts of Treatment Setup Error

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Yang, L; Southern Medical University, Guangzhou; Tian, Z

    2014-06-01

    Purpose: Current Intensity Modulated Radiation Therapy (IMRT) is plan-centered. At each treatment fraction, we position the patient to match the setup in treatment plan. Inaccurate setup can compromise delivered dose distribution, and hence leading to suboptimal treatments. Moreover, current setup approach via couch shift under image guidance can correct translational errors, while rotational and deformation errors are hard to address. To overcome these problems, we propose in this abstract a patient-centered scheme to mitigate impacts of treatment setup errors. Methods: In the patient-centered scheme, we first position the patient on the couch approximately matching the planned-setup. Our Supercomputing Online Replanningmore » Environment (SCORE) is then employed to design an optimal treatment plan based on the daily patient geometry. It hence mitigates the impacts of treatment setup error and reduces the requirements on setup accuracy. We have conducted simulations studies in 10 head-and-neck (HN) patients to investigate the feasibility of this scheme. Rotational and deformation setup errors were simulated. Specifically, 1, 3, 5, 7 degrees of rotations were put on pitch, roll, and yaw directions; deformation errors were simulated by splitting neck movements into four basic types: rotation, lateral bending, flexion and extension. Setup variation ranges are based on observed numbers in previous studies. Dosimetric impacts of our scheme were evaluated on PTVs and OARs in comparison with original plan dose with original geometry and original plan recalculated dose with new setup geometries. Results: With conventional plan-centered approach, setup error could lead to significant PTV D99 decrease (−0.25∼+32.42%) and contralateral-parotid Dmean increase (−35.09∼+42.90%). The patientcentered approach is effective in mitigating such impacts to 0∼+0.20% and −0.03∼+5.01%, respectively. Computation time is <128 s. Conclusion: Patient-centered scheme is proposed to mitigate setup error impacts using replanning. Its superiority in terms of dosimetric impacts and feasibility has been shown through simulation studies on HN cases.« less

  1. Design, analysis, and test verification of advanced encapsulation systems

    NASA Technical Reports Server (NTRS)

    Mardesich, N.; Minning, C.

    1982-01-01

    Design sensitivities are established for the development of photovoltaic module criteria and the definition of needed research tasks. The program consists of three phases. In Phase I, analytical models were developed to perform optical, thermal, electrical, and structural analyses on candidate encapsulation systems. From these analyses several candidate systems will be selected for qualification testing during Phase II. Additionally, during Phase II, test specimens of various types will be constructed and tested to determine the validity of the analysis methodology developed in Phase I. In Phse III, a finalized optimum design based on knowledge gained in Phase I and II will be developed. All verification testing was completed during this period. Preliminary results and observations are discussed. Descriptions of the thermal, thermal structural, and structural deflection test setups are included.

  2. Constitutive modeling of superalloy single crystals with verification testing

    NASA Technical Reports Server (NTRS)

    Jordan, Eric; Walker, Kevin P.

    1985-01-01

    The goal is the development of constitutive equations to describe the elevated temperature stress-strain behavior of single crystal turbine blade alloys. The program includes both the development of a suitable model and verification of the model through elevated temperature-torsion testing. A constitutive model is derived from postulated constitutive behavior on individual crystallographic slip systems. The behavior of the entire single crystal is then arrived at by summing up the slip on all the operative crystallographic slip systems. This type of formulation has a number of important advantages, including the prediction orientation dependence and the ability to directly represent the constitutive behavior in terms which metallurgists use in describing the micromechanisms. Here, the model is briefly described, followed by the experimental set-up and some experimental findings to date.

  3. Chapter 51: How to Build a Simple Cone Search Service Using a Local Database

    NASA Astrophysics Data System (ADS)

    Kent, B. R.; Greene, G. R.

    The cone search service protocol will be examined from the server side in this chapter. A simple cone search service will be setup and configured locally using MySQL. Data will be read into a table, and the Java JDBC will be used to connect to the database. Readers will understand the VO cone search specification and how to use it to query a database on their local systems and return an XML/VOTable file based on an input of RA/DEC coordinates and a search radius. The cone search in this example will be deployed as a Java servlet. The resulting cone search can be tested with a verification service. This basic setup can be used with other languages and relational databases.

  4. [Statistical Process Control (SPC) can help prevent treatment errors without increasing costs in radiotherapy].

    PubMed

    Govindarajan, R; Llueguera, E; Melero, A; Molero, J; Soler, N; Rueda, C; Paradinas, C

    2010-01-01

    Statistical Process Control (SPC) was applied to monitor patient set-up in radiotherapy and, when the measured set-up error values indicated a loss of process stability, its root cause was identified and eliminated to prevent set-up errors. Set up errors were measured for medial-lateral (ml), cranial-caudal (cc) and anterior-posterior (ap) dimensions and then the upper control limits were calculated. Once the control limits were known and the range variability was acceptable, treatment set-up errors were monitored using sub-groups of 3 patients, three times each shift. These values were plotted on a control chart in real time. Control limit values showed that the existing variation was acceptable. Set-up errors, measured and plotted on a X chart, helped monitor the set-up process stability and, if and when the stability was lost, treatment was interrupted, the particular cause responsible for the non-random pattern was identified and corrective action was taken before proceeding with the treatment. SPC protocol focuses on controlling the variability due to assignable cause instead of focusing on patient-to-patient variability which normally does not exist. Compared to weekly sampling of set-up error in each and every patient, which may only ensure that just those sampled sessions were set-up correctly, the SPC method enables set-up error prevention in all treatment sessions for all patients and, at the same time, reduces the control costs. Copyright © 2009 SECA. Published by Elsevier Espana. All rights reserved.

  5. Machine-assisted verification of latent fingerprints: first results for nondestructive contact-less optical acquisition techniques with a CWL sensor

    NASA Astrophysics Data System (ADS)

    Hildebrandt, Mario; Kiltz, Stefan; Krapyvskyy, Dmytro; Dittmann, Jana; Vielhauer, Claus; Leich, Marcus

    2011-11-01

    A machine-assisted analysis of traces from crime scenes might be possible with the advent of new high-resolution non-destructive contact-less acquisition techniques for latent fingerprints. This requires reliable techniques for the automatic extraction of fingerprint features from latent and exemplar fingerprints for matching purposes using pattern recognition approaches. Therefore, we evaluate the NIST Biometric Image Software for the feature extraction and verification of contact-lessly acquired latent fingerprints to determine potential error rates. Our exemplary test setup includes 30 latent fingerprints from 5 people in two test sets that are acquired from different surfaces using a chromatic white light sensor. The first test set includes 20 fingerprints on two different surfaces. It is used to determine the feature extraction performance. The second test set includes one latent fingerprint on 10 different surfaces and an exemplar fingerprint to determine the verification performance. This utilized sensing technique does not require a physical or chemical visibility enhancement of the fingerprint residue, thus the original trace remains unaltered for further investigations. No particular feature extraction and verification techniques have been applied to such data, yet. Hence, we see the need for appropriate algorithms that are suitable to support forensic investigations.

  6. Verification of the SENTINEL-4 Focal Plane Subsystem

    NASA Astrophysics Data System (ADS)

    Williges, C.; Hohn, R.; Rossmann, H.; Hilbert, S.; Uhlig, M.; Buchwinkler, K.; Reulke, R.

    2017-05-01

    The Sentinel-4 payload is a multi-spectral camera system which is designed to monitor atmospheric conditions over Europe. The German Aerospace Center (DLR) in Berlin, Germany conducted the verification campaign of the Focal Plane Subsystem (FPS) on behalf of Airbus Defense and Space GmbH, Ottobrunn, Germany. The FPS consists, inter alia, of two Focal Plane Assemblies (FPAs), one for the UV-VIS spectral range (305 nm … 500 nm), the second for NIR (750 nm … 775 nm). In this publication, we will present in detail the opto-mechanical laboratory set-up of the verification campaign of the Sentinel-4 Qualification Model (QM) which will also be used for the upcoming Flight Model (FM) verification. The test campaign consists mainly of radiometric tests performed with an integrating sphere as homogenous light source. The FPAs have mainly to be operated at 215 K ± 5 K, making it necessary to exploit a thermal vacuum chamber (TVC) for the test accomplishment. This publication focuses on the challenge to remotely illuminate both Sentinel-4 detectors as well as a reference detector homogeneously over a distance of approximately 1 m from outside the TVC. Furthermore selected test analyses and results will be presented, showing that the Sentinel-4 FPS meets specifications.

  7. A case study for online plan adaptation using helical tomotherapy

    PubMed Central

    Neilson, Christopher E.; Yartsev, Slav

    2012-01-01

    Helical tomotherapy's ability to provide daily megavoltage (MV) computed tomography (CT) images for patient set-up verification allows for the creation of adapted plans. As plans become more complex by introducing sharper dose gradients in an effort to spare healthy tissue, inter-fraction changes of organ position with respect to plan become a limiting factor in the correct dose delivery to the target. Tomotherapy's planned adaptive option provides the possibility to evaluate the dose distribution for each fraction and subsequently adapt the original plan to the current anatomy. In this study, 30 adapted plans were created using new contours based on the daily MVCT studies of a bladder cancer patient with considerable anatomical variations. Dose to the rectum and two planning target volumes (PTVs) were compared between the original plan, the dose that was actually delivered to the patient, and the theoretical dose from the 30 adapted plans. The adaptation simulation displayed a lower dose to 35% and 50% of the rectum compared to no adaptation at all, while maintaining an equivalent dose to the PTVs. Although online adaptation is currently too time-consuming, it has the potential to improve the effectiveness of radiotherapy. PMID:22557799

  8. In vivo verification of particle therapy: how Compton camera configurations affect 3D image quality

    NASA Astrophysics Data System (ADS)

    Mackin, D.; Draeger, E.; Peterson, S.; Polf, J.; Beddar, S.

    2017-05-01

    The steep dose gradients enabled by the Bragg peaks of particle therapy beams are a double edged sword. They enable highly conformal dose distributions, but even small deviations from the planned beam range can cause overdosing of healthy tissue or under-dosing of the tumour. To reduce this risk, particle therapy treatment plans include margins large enough to account for all the sources of range uncertainty, which include patient setup errors, patient anatomy changes, and CT number to stopping power ratios. Any system that could verify the beam range in vivo, would allow reduced margins and more conformal dose distributions. Toward our goal developing such a system based on Compton camera (CC) imaging, we studied how three configurations (single camera, parallel opposed, and orthogonal) affect the quality of the 3D images. We found that single CC and parallel opposed configurations produced superior images in 2D. The increase in parallax produced by an orthogonal CC configuration was shown to be beneficial in producing artefact free 3D images.

  9. Integration of Error Compensation of Coordinate Measuring Machines into Feature Measurement: Part II—Experimental Implementation

    PubMed Central

    Calvo, Roque; D’Amato, Roberto; Gómez, Emilio; Domingo, Rosario

    2016-01-01

    Coordinate measuring machines (CMM) are main instruments of measurement in laboratories and in industrial quality control. A compensation error model has been formulated (Part I). It integrates error and uncertainty in the feature measurement model. Experimental implementation for the verification of this model is carried out based on the direct testing on a moving bridge CMM. The regression results by axis are quantified and compared to CMM indication with respect to the assigned values of the measurand. Next, testing of selected measurements of length, flatness, dihedral angle, and roundness features are accomplished. The measurement of calibrated gauge blocks for length or angle, flatness verification of the CMM granite table and roundness of a precision glass hemisphere are presented under a setup of repeatability conditions. The results are analysed and compared with alternative methods of estimation. The overall performance of the model is endorsed through experimental verification, as well as the practical use and the model capability to contribute in the improvement of current standard CMM measuring capabilities. PMID:27754441

  10. Setup errors and effectiveness of Optical Laser 3D Surface imaging system (Sentinel) in postoperative radiotherapy of breast cancer.

    PubMed

    Wei, Xiaobo; Liu, Mengjiao; Ding, Yun; Li, Qilin; Cheng, Changhai; Zong, Xian; Yin, Wenming; Chen, Jie; Gu, Wendong

    2018-05-08

    Breast-conserving surgery (BCS) plus postoperative radiotherapy has become the standard treatment for early-stage breast cancer. The aim of this study was to compare the setup accuracy of optical surface imaging by the Sentinel system with cone-beam computerized tomography (CBCT) imaging currently used in our clinic for patients received BCS. Two optical surface scans were acquired before and immediately after couch movement correction. The correlation between the setup errors as determined by the initial optical surface scan and CBCT was analyzed. The deviation of the second optical surface scan from the reference planning CT was considered an estimate for the residual errors for the new method for patient setup correction. The consequences in terms for necessary planning target volume (PTV) margins for treatment sessions without setup correction applied. We analyzed 145 scans in 27 patients treated for early stage breast cancer. The setup errors of skin marker based patient alignment by optical surface scan and CBCT were correlated, and the residual setup errors as determined by the optical surface scan after couch movement correction were reduced. Optical surface imaging provides a convenient method for improving the setup accuracy for breast cancer patient without unnecessary imaging dose.

  11. Experimental verification of Pyragas-Schöll-Fiedler control.

    PubMed

    von Loewenich, Clemens; Benner, Hartmut; Just, Wolfram

    2010-09-01

    We present an experimental realization of time-delayed feedback control proposed by Schöll and Fiedler. The scheme enables us to stabilize torsion-free periodic orbits in autonomous systems, and to overcome the so-called odd number limitation. The experimental control performance is in quantitative agreement with the bifurcation analysis of simple model systems. The results uncover some general features of the control scheme which are deemed to be relevant for a large class of setups.

  12. Experimental verification of PSM polarimetry: monitoring polarization at 193nm high-NA with phase shift masks

    NASA Astrophysics Data System (ADS)

    McIntyre, Gregory; Neureuther, Andrew; Slonaker, Steve; Vellanki, Venu; Reynolds, Patrick

    2006-03-01

    The initial experimental verification of a polarization monitoring technique is presented. A series of phase shifting mask patterns produce polarization dependent signals in photoresist and are capable of monitoring the Stokes parameters of any arbitrary illumination scheme. Experiments on two test reticles have been conducted. The first reticle consisted of a series of radial phase gratings (RPG) and employed special apertures to select particular illumination angles. Measurement sensitivities of about 0.3 percent of the clear field per percent change in polarization state were observed. The second test reticle employed the more sensitive proximity effect polarization analyzers (PEPA), a more robust experimental setup, and a backside pinhole layer for illumination angle selection and to enable characterization of the full illuminator. Despite an initial complication with the backside pinhole alignment, the results correlate with theory. Theory suggests that, once the pinhole alignment is corrected in the near future, the second reticle should achieve a measurement sensitivity of about 1 percent of the clear field per percent change in polarization state. This corresponds to a measurement of the Stokes parameters after test mask calibration, to within about 0.02 to 0.03. Various potential improvements to the design, fabrication of the mask, and experimental setup are discussed. Additionally, to decrease measurement time, a design modification and double exposure technique is proposed to enable electrical detection of the measurement signal.

  13. Design and development of new collimator cones for fractionated stereotactic radiation therapy in Samsung Medical Center.

    PubMed

    Ahn, Y C; Ju, S G; Kim, D Y; Choi, D R; Huh, S J; Park, Y H; Lim, D H; Kim, M K

    1999-05-01

    In stereotactic radiotherapy using X-Knife system, the commercially supplied collimator cone system had a few mechanical limitations. The authors have developed new collimator cones to overcome these limitations and named them "SMC type" collimator cones. We made use of cadmium-free cerrobend alloy within the stainless steel cylinder housing. We made nine cones of relatively larger sizes (3.0 cm to 7.0 cm in diameter) and of shorter length with bigger clearance from the isocenter than the commercial cones. The cone housing and the collimator cones were designed to insert into the wedge mount of the gantry head to enable double-exposure linac-gram taking. The mechanical accuracy of pointing to the isocenter was tested by ball test and cone rotation test, and the dosimetric measurements were performed, all of which were with satisfactory results. A new innovative quality assurance procedure using linac-grams on the patients at the actual treatment setup was attempted after taking 10 sets of AP and lateral linac-grams and the overall mechanical isocenter accuracy was excellent (average error = 0.4 +/- 0.2 mm). We have developed the SMC type collimator cone system mainly for fractionated stereotactic radiation therapy use with our innovative ideas. The new cones' mechanical accuracy and physical properties were satisfactory for clinical use, and the verification of the isocenter accuracy on the actual treatment setup has become possible.

  14. SU-E-T-424: Dosimetric Verification of Modulated Electron Radiation Therapy Delivered Using An Electron Specific Multileaf Collimator for Treatment of Scalp Cases

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Eldib, A; Al-Azhar University Cairo; Jin, L

    2014-06-01

    Purpose: Modulated electron radiotherapy (MERT) has the potential to achieve better treatment outcome for shallow tumors such as those of breast and scalp. In a separate study with scalp lesions, MERT was compared to volumetric modulated arc therapy. Our results showed a reduction in the dose reaching the brain with MERT. However dose calculation accuracy and delivery efficiency challenges remain. Thus in the current study we proceed to add more cases to demonstrate MERT beneficial outcome and its delivery accuracy using an electron specific multileaf collimator (eMLC). Methods: We have used the MCBEAM code for treatment head simulation and formore » generating phase space files to be used as radiation source input for our Monte Carlo based treatment planning system (MC TPS). MCPLAN code is used for calculation of patient specific dose deposition coefficient and for final MERT plan dose calculation. An in-house developed optimization code is used for the optimization process. MERT plans were generated for real patients and head and neck phantom. Film was used for dosimetric verification. The film was cut following the contour of the curved phantom surface and then sealed with black masking tape. In the measurement, the sealed film packet was sandwiched between two adjacent slabs of the head and neck phantom. The measured 2D dose distribution was then compared with calculations. Results: The eMLC allows effective treatment of scalps with multi-lesions spreading around the patient head, which was usually difficult to plan or very time consuming with conventional applicators. MERT continues to show better reduction in the brain dose. The dosimetric measurements showed slight discrepancy, which was attributed to the film setup. Conclusion: MERT can improve treatment plan quality for patients with scalp cancers. Our in-house MC TPS is capable of performing treatment planning and accurate dose calculation for MERT using the eMLC.« less

  15. Automatization of hardware configuration for plasma diagnostic system

    NASA Astrophysics Data System (ADS)

    Wojenski, A.; Pozniak, K. T.; Kasprowicz, G.; Kolasinski, P.; Krawczyk, R. D.; Zabolotny, W.; Linczuk, P.; Chernyshova, M.; Czarski, T.; Malinowski, K.

    2016-09-01

    Soft X-ray plasma measurement systems are mostly multi-channel, high performance systems. In case of the modular construction it is necessary to perform sophisticated system discovery in parallel with automatic system configuration. In the paper the structure of the modular system designed for tokamak plasma soft X-ray measurements is described. The concept of the system discovery and further automatic configuration is also presented. FCS application (FMC/ FPGA Configuration Software) is used for running sophisticated system setup with automatic verification of proper configuration. In order to provide flexibility of further system configurations (e.g. user setup), common communication interface is also described. The approach presented here is related to the automatic system firmware building presented in previous papers. Modular construction and multichannel measurements are key requirement in term of SXR diagnostics with use of GEM detectors.

  16. SU-E-T-235: Data Mining for Evaluating Treatment Performances Over a Large Quantity of Data to Monitor and Improve SBRT Workflow

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Feng, W; Bayhealth Medical Center, Dover, DE; Chu, A

    Purpose: To quality assure a large quantity of retrospective treatment cases for treatment performances by randomly sampling is inefficient. Here we provide a method to efficiently monitor and investigate the QA of SBRT workflow over Mosaiq. Methods: The code developed with Microsoft SQL Server Management Studio 2008R2 and VBA was used for retrieving and sorting data from Mosaiq (version 2.3–2.6 during 2012–2015). SBRT patients were filtered by fractional dose over 350cGy and total fraction number less than 6, which SBRT prescriptions were defined. The quality assurance on the SBRT workflow was focused on the treatment deliveries such as patient positioningmore » setup, CBCT indicated offsets and couch shifted corrections. The treatment delivery were done by Varian Truebeam systems and the record/verify by Mosaiq. Results: Total 82 SBRT patients corresponding to 103 courses and 854 CBCT images were found by the retrieval query. Most centers record daily pre-treatment (Pre-Tx: before treatment shift) image-guided shifts along treatment course for inter-fraction motion record, and it is useful to also verify it with post-treatment imaging (Post-Tx: after treatment CBCT verification) to verify intra-fraction motion. Analyzing the details of daily recorded shifts can reveals the information of patient-setup and staff’s record/verify behaviors. 3 examples were provided as solid evidences and on-going rectification for preventing future mistakes. Conclusions: The report gave feasible examples for inspector to verify a large amount of data during site investigation. This program can also be extended to a scheduled data mining with software to periodical analyze the timely records in Mosaiq, for example, a various control charts for different QA purposes. As the current trend of automation in radiation therapy field, the data mining would be a necessary tool in the future, just as the automatic plan quality evaluation has been under development in Eclipse.« less

  17. SU-E-J-55: End-To-End Effectiveness Analysis of 3D Surface Image Guided Voluntary Breath-Holding Radiotherapy for Left Breast

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lin, M; Feigenberg, S

    Purpose To evaluate the effectiveness of using 3D-surface-image to guide breath-holding (BH) left-side breast treatment. Methods Two 3D surface image guided BH procedures were implemented and evaluated: normal-BH, taking BH at a comfortable level, and deep-inspiration-breath-holding (DIBH). A total of 20 patients (10 Normal-BH and 10 DIBH) were recruited. Patients received a BH evaluation using a commercialized 3D-surface- tracking-system (VisionRT, London, UK) to quantify the reproducibility of BH positions prior to CT scan. Tangential 3D/IMRT plans were conducted. Patients were initially setup under free-breathing (FB) condition using the FB surface obtained from the untaged CT to ensure a correct patientmore » position. Patients were then guided to reach the planned BH position using the BH surface obtained from the BH CT. Action-levels were set at each phase of treatment process based on the information provided by the 3D-surface-tracking-system for proper interventions (eliminate/re-setup/ re-coaching). We reviewed the frequency of interventions to evaluate its effectiveness. The FB-CBCT and port-film were utilized to evaluate the accuracy of 3D-surface-guided setups. Results 25% of BH candidates with BH positioning uncertainty > 2mm are eliminated prior to CT scan. For >90% of fractions, based on the setup deltas from3D-surface-trackingsystem, adjustments of patient setup are needed after the initial-setup using laser. 3D-surface-guided-setup accuracy is comparable as CBCT. For the BH guidance, frequency of interventions (a re-coaching/re-setup) is 40%(Normal-BH)/91%(DIBH) of treatments for the first 5-fractions and then drops to 16%(Normal-BH)/46%(DIBH). The necessity of re-setup is highly patient-specific for Normal-BH but highly random among patients for DIBH. Overall, a −0.8±2.4 mm accuracy of the anterior pericardial shadow position was achieved. Conclusion 3D-surface-image technology provides effective intervention to the treatment process and ensures favorable day-to-day setup accuracy. DIBH setup appears to be more uncertain and this would be the patient group who will definitely benefit from the extra information of 3D surface setup.« less

  18. Effects of megavoltage computed tomographic scan methodology on setup verification and adaptive dose calculation in helical TomoTherapy.

    PubMed

    Zhu, Jian; Bai, Tong; Gu, Jiabing; Sun, Ziwen; Wei, Yumei; Li, Baosheng; Yin, Yong

    2018-04-27

    To evaluate the effect of pretreatment megavoltage computed tomographic (MVCT) scan methodology on setup verification and adaptive dose calculation in helical TomoTherapy. Both anthropomorphic heterogeneous chest and pelvic phantoms were planned with virtual targets by TomoTherapy Physicist Station and were scanned with TomoTherapy megavoltage image-guided radiotherapy (IGRT) system consisted of six groups of options: three different acquisition pitches (APs) of 'fine', 'normal' and 'coarse' were implemented by multiplying 2 different corresponding reconstruction intervals (RIs). In order to mimic patient setup variations, each phantom was shifted 5 mm away manually in three orthogonal directions respectively. The effect of MVCT scan options was analyzed in image quality (CT number and noise), adaptive dose calculation deviations and positional correction variations. MVCT scanning time with pitch of 'fine' was approximately twice of 'normal' and 3 times more than 'coarse' setting, all which will not be affected by different RIs. MVCT with different APs delivered almost identical CT numbers and image noise inside 7 selected regions with various densities. DVH curves from adaptive dose calculation with serial MVCT images acquired by varied pitches overlapped together, where as there are no significant difference in all p values of intercept & slope of emulational spinal cord (p = 0.761 & 0.277), heart (p = 0.984 & 0.978), lungs (p = 0.992 & 0.980), soft tissue (p = 0.319 & 0.951) and bony structures (p = 0.960 & 0.929) between the most elaborated and the roughest serials of MVCT. Furthermore, gamma index analysis shown that, compared to the dose distribution calculated on MVCT of 'fine', only 0.2% or 1.1% of the points analyzed on MVCT of 'normal' or 'coarse' do not meet the defined gamma criterion. On chest phantom, all registration errors larger than 1 mm appeared at superior-inferior axis, which cannot be avoided with the smallest AP and RI. On pelvic phantom, craniocaudal errors are much smaller than chest, however, AP of 'coarse' presents larger registration errors which can be reduced from 2.90 mm to 0.22 mm by registration technique of 'full image'. AP of 'coarse' with RI of 6 mm is recommended in adaptive radiotherapy (ART) planning to provide craniocaudal longer and faster MVCT scan, while registration technique of 'full image' should be used to avoid large residual error. Considering the trade-off between IGRT and ART, AP of 'normal' with RI of 2 mm was highly recommended in daily practice.

  19. Verification of the proteus two-dimensional Navier-Stokes code for flat plate and pipe flows

    NASA Technical Reports Server (NTRS)

    Conley, Julianne M.; Zeman, Patrick L.

    1991-01-01

    The Proteus Navier-Stokes Code is evaluated for 2-D/axisymmetric, viscous, incompressible, internal, and external flows. The particular cases to be discussed are laminar and turbulent flows over a flat plate, laminar and turbulent developing pipe flows, and turbulent pipe flow with swirl. Results are compared with exact solutions, empirical correlations, and experimental data. A detailed description of the code set-up, including boundary conditions, initial conditions, grid size, and grid packing is given for each case.

  20. Demonstration of measurement-only blind quantum computing

    NASA Astrophysics Data System (ADS)

    Greganti, Chiara; Roehsner, Marie-Christine; Barz, Stefanie; Morimae, Tomoyuki; Walther, Philip

    2016-01-01

    Blind quantum computing allows for secure cloud networks of quasi-classical clients and a fully fledged quantum server. Recently, a new protocol has been proposed, which requires a client to perform only measurements. We demonstrate a proof-of-principle implementation of this measurement-only blind quantum computing, exploiting a photonic setup to generate four-qubit cluster states for computation and verification. Feasible technological requirements for the client and the device-independent blindness make this scheme very applicable for future secure quantum networks.

  1. Hawking radiation in an electromagnetic waveguide?

    PubMed

    Schützhold, Ralf; Unruh, William G

    2005-07-15

    It is demonstrated that the propagation of electromagnetic waves in an appropriately designed waveguide is (for large wavelengths) analogous to that within a curved space-time--such as around a black hole. As electromagnetic radiation (e.g., microwaves) can be controlled, amplified, and detected (with present-day technology) much easier than sound, for example, we propose a setup for the experimental verification of the Hawking effect. Apart from experimentally testing this striking prediction, this would facilitate the investigation of the trans-Planckian problem.

  2. The plenoptic camera as a wavefront sensor for the European Solar Telescope (EST)

    NASA Astrophysics Data System (ADS)

    Rodríguez-Ramos, Luis F.; Martín, Yolanda; Díaz, José J.; Piqueras, J.; Rodríguez-Ramos, J. M.

    2009-08-01

    The plenoptic wavefront sensor combines measurements at pupil and image planes in order to obtain wavefront information from different points of view simultaneously, being capable to sample the volume above the telescope to extract the tomographic information of the atmospheric turbulence. After describing the working principle, a laboratory setup has been used for the verification of the capability of measuring the pupil plane wavefront. A comparative discussion with respect to other wavefront sensors is also included.

  3. Automated patient identification and localization error detection using 2-dimensional to 3-dimensional registration of kilovoltage x-ray setup images.

    PubMed

    Lamb, James M; Agazaryan, Nzhde; Low, Daniel A

    2013-10-01

    To determine whether kilovoltage x-ray projection radiation therapy setup images could be used to perform patient identification and detect gross errors in patient setup using a computer algorithm. Three patient cohorts treated using a commercially available image guided radiation therapy (IGRT) system that uses 2-dimensional to 3-dimensional (2D-3D) image registration were retrospectively analyzed: a group of 100 cranial radiation therapy patients, a group of 100 prostate cancer patients, and a group of 83 patients treated for spinal lesions. The setup images were acquired using fixed in-room kilovoltage imaging systems. In the prostate and cranial patient groups, localizations using image registration were performed between computed tomography (CT) simulation images from radiation therapy planning and setup x-ray images corresponding both to the same patient and to different patients. For the spinal patients, localizations were performed to the correct vertebral body, and to an adjacent vertebral body, using planning CTs and setup x-ray images from the same patient. An image similarity measure used by the IGRT system image registration algorithm was extracted from the IGRT system log files and evaluated as a discriminant for error detection. A threshold value of the similarity measure could be chosen to separate correct and incorrect patient matches and correct and incorrect vertebral body localizations with excellent accuracy for these patient cohorts. A 10-fold cross-validation using linear discriminant analysis yielded misclassification probabilities of 0.000, 0.0045, and 0.014 for the cranial, prostate, and spinal cases, respectively. An automated measure of the image similarity between x-ray setup images and corresponding planning CT images could be used to perform automated patient identification and detection of localization errors in radiation therapy treatments. Copyright © 2013 Elsevier Inc. All rights reserved.

  4. Helical tomotherapy setup variations in canine nasal tumor patients immobilized with a bite block.

    PubMed

    Kubicek, Lyndsay N; Seo, Songwon; Chappell, Richard J; Jeraj, Robert; Forrest, Lisa J

    2012-01-01

    The purpose of our study was to compare setup variation in four degrees of freedom (vertical, longitudinal, lateral, and roll) between canine nasal tumor patients immobilized with a mattress and bite block, versus a mattress alone. Our secondary aim was to define a clinical target volume (CTV) to planning target volume (PTV) expansion margin based on our mean systematic error values associated with nasal tumor patients immobilized by a mattress and bite block. We evaluated six parameters for setup corrections: systematic error, random error, patient-patient variation in systematic errors, the magnitude of patient-specific random errors (root mean square [RMS]), distance error, and the variation of setup corrections from zero shift. The variations in all parameters were statistically smaller in the group immobilized by a mattress and bite block. The mean setup corrections in the mattress and bite block group ranged from 0.91 mm to 1.59 mm for the translational errors and 0.5°. Although most veterinary radiation facilities do not have access to Image-guided radiotherapy (IGRT), we identified a need for more rigid fixation, established the value of adding IGRT to veterinary radiation therapy, and define the CTV-PTV setup error margin for canine nasal tumor patients immobilized in a mattress and bite block. © 2012 Veterinary Radiology & Ultrasound.

  5. Physical aspects of total-body irradiation at the Middlesex Hospital (UCL group of hospitals), London 1988-1993: I. Phantom measurements and planning methods.

    PubMed

    Planskoy, B; Bedford, A M; Davis, F M; Tapper, P D; Loverock, L T

    1996-11-01

    This paper, which is divided into parts I and II, describes the physical aspects of work on total-body irradiation (TBI) at the Middlesex Hospital, London, from 1988 to 1993. Irradiation is fractionated and bi-lateral with horizontal accelerator photon beams of 8 MV (1988-1992) at a source-surface distance (SSD) of 3.36 m and 10 MV (1992-1993) at an SSD of 4.62 m. The main aims were maximum patient comfort, a simple, accurate set-up with overall times per fraction of 30 min or less, dose homogeneity throughout the body within +/- 10 to +/- 15%, pre-irradiation treatment planning on nine CT slices using our commercial IGE RTplan (1988-1992) and Target 2 (1992-1993) treatment planning systems and, most important, verification of the plans by in vivo dosimetry to within +/- 5%. Verification of the planned lung doses, which are distributed over five CT slices, was given special attention. In part I of this paper we describe the preliminary work, most of which was done prior to patient treatment. This consisted of standard dosimetric measurements (central axis depth doses, beam profiles at several depths, build-up and build-down curves, beam output calibrations, effect of body compensators, etc), in evaluating silicon diode dosimeters for in vivo dosimetry and of adapting and verifying the methods of treatment planning for TBI conditions. The results obtained with phantoms, including a Rando body phantom, showed that, in principle, our aims could be achieved. The final proof depended, however, on an analysis of the results of the in vivo work and this forms the subject of part II of this paper.

  6. A double exposed portal image comparison between electronic portal imaging hard copies and port films in radiation therapy treatment setup confirmation to determine its clinical application in a radiotherapy center.

    PubMed

    Hatherly, K E; Smylie, J C; Rodger, A; Dally, M J; Davis, S R; Millar, J L

    2001-01-01

    At the William Buckland Radiotherapy Center (WBRC), field-only electronic portal image (EPI) hard copies are used for radiation treatment field verification for whole brain, breast, chest, spine, and large pelvic fields, as determined by a previous study. A subsequent research project, addressing the quality of double exposed EPI hard copies for sites where field only EPI was not considered adequate to determine field placement, has been undertaken. The double exposed EPI hard copies were compared to conventional double exposed port films for small pelvic, partial brain, and head and neck fields and for a miscellaneous group. All double exposed EPIs were captured during routine clinical procedures using liquid ion chamber cassettes. EPI hard copies were generated using a Visiplex multi-format camera. In sites where port film remained the preferred verification format, the port films were generated as per department protocol. In addition EPIs were collected specifically for this project. Four radiation oncologists performed the evaluation of EPI and port film images independently with a questionnaire completed at each stage of the evaluation process to assess the following: Adequacy of information in the image to assess field placement. Adequacy of information for determining field placement correction. Clinician's preferred choice of imaging for field placement assessment The results indicate that double exposed EPI hard copies generally do containsufficient information to permit evaluation of field placement and can replace conventionaldouble exposed port films in a significant number of sites. These include the following:pelvis fields < 12 X 12 cm, partial brain fields, and a miscellaneous group. However forradical head and neck fields, the preferred verification image format remained port film dueto the image hard copy size and improved contrast for this media. Thus in this departmenthard copy EPI is the preferred modality of field verification for all sites except radical headand neck treatments. This should result in an increase in efficiency of workloadmanagement and patient care.

  7. Integrated Advanced Microwave Sounding Unit-A (AMSU-A). Performance Verification Report: METSAT (S/N 108) AMSU-A1 Receiver Assemblies, P/N 1356429-1 S/N F05 and P/N 1356409-1 S/N F05

    NASA Technical Reports Server (NTRS)

    Haigh, R.; Krimchansky, S. (Technical Monitor)

    2000-01-01

    This is the Performance Verification Report, METSAT (S/N 108) AMSU-A1 Receiver Assemblies P/N 1356429-1 S/N F05 and P/N 1356409-1 S/N F05, for the Integrated Advanced Microwave Sounding Unit-A (AMSU-A). The ATP for the AMSU-A Receiver Subsystem, AE-26002/6A, is prepared to describe in detail the configuration of the test setups and the procedures of the tests to verify that the receiver subsystem meets the specifications as required either in the AMSU-A Instrument Performance and Operation Specifications, S-480-80, or in AMSU-A Receiver Subsystem Specifications, AE-26608, derived by the Aerojet System Engineering. Test results that verify the conformance to the specifications demonstrate the acceptability of that particular receiver subsystem.

  8. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Tanabe, S; Utsunomiya, S; Abe, E

    Purpose: To assess an accuracy of fiducial maker-based setup using ExacTrac (ExT-based setup) as compared with soft tissue-based setup using Cone-beam CT (CBCT-based setup) for patients with prostate cancer receiving intensity-modulated radiation therapy (IMRT) for the purpose of investigating whether ExT-based setup can be an alternative to CBCT-based setup. Methods: The setup accuracy was analyzed prospectively for 7 prostate cancer patients with implanted three fiducial markers received IMRT. All patients were treated after CBCT-based setup was performed and corresponding shifts were recorded. ExacTrac images were obtained before and after CBCT-based setup. The fiducial marker-based shifts were calculated based on thosemore » two images and recorded on the assumption that the setup correction was carried out by fiducial marker-based auto correction. Mean and standard deviation of absolute differences and the correlation between CBCT and ExT shifts were estimated. Results: A total of 178 image dataset were analyzed. On the differences between CBCT and ExT shifts, 133 (75%) of 178 image dataset resulted in smaller differences than 3 mm in all dimensions. Mean differences in the anterior-posterior (AP), superior-inferior (SI), and left-right (LR) dimensions were 1.8 ± 1.9 mm, 0.7 ± 1.9 mm, and 0.6 ± 0.8 mm, respectively. The percentages of shift agreements within ±3 mm were 76% for AP, 90% for SI, and 100% for LR. The Pearson coefficient of correlation for CBCT and ExT shifts were 0.80 for AP, 0.80 for SI, and 0.65 for LR. Conclusion: This work showed that the accuracy of ExT-based setup was correlated with that of CBCT-based setup, implying that ExT-based setup has a potential ability to be an alternative to CBCT-based setup. The further work is to specify the conditions that ExT-based setup can provide the accuracy comparable to CBCT-based setup.« less

  9. Cone beam CT-based set-up strategies with and without rotational correction for stereotactic body radiation therapy in the liver.

    PubMed

    Bertholet, Jenny; Worm, Esben; Høyer, Morten; Poulsen, Per

    2017-06-01

    Accurate patient positioning is crucial in stereotactic body radiation therapy (SBRT) due to a high dose regimen. Cone-beam computed tomography (CBCT) is often used for patient positioning based on radio-opaque markers. We compared six CBCT-based set-up strategies with or without rotational correction. Twenty-nine patients with three implanted markers received 3-6 fraction liver SBRT. The markers were delineated on the mid-ventilation phase of a 4D-planning-CT. One pretreatment CBCT was acquired per fraction. Set-up strategy 1 used only translational correction based on manual marker match between the CBCT and planning CT. Set-up strategy 2 used automatic 6 degrees-of-freedom registration of the vertebrae closest to the target. The 3D marker trajectories were also extracted from the projections and the mean position of each marker was calculated and used for set-up strategies 3-6. Translational correction only was used for strategy 3. Translational and rotational corrections were used for strategies 4-6 with the rotation being either vertebrae based (strategy 4), or marker based and constrained to ±3° (strategy 5) or unconstrained (strategy 6). The resulting set-up error was calculated as the 3D root-mean-square set-up error of the three markers. The set-up error of the spinal cord was calculated for all strategies. The bony anatomy set-up (2) had the largest set-up error (5.8 mm). The marker-based set-up with unconstrained rotations (6) had the smallest set-up error (0.8 mm) but the largest spinal cord set-up error (12.1 mm). The marker-based set-up with translational correction only (3) or with bony anatomy rotational correction (4) had equivalent set-up error (1.3 mm) but rotational correction reduced the spinal cord set-up error from 4.1 mm to 3.5 mm. Marker-based set-up was substantially better than bony-anatomy set-up. Rotational correction may improve the set-up, but further investigations are required to determine the optimal correction strategy.

  10. Multivariate statistical monitoring as applied to clean-in-place (CIP) and steam-in-place (SIP) operations in biopharmaceutical manufacturing.

    PubMed

    Roy, Kevin; Undey, Cenk; Mistretta, Thomas; Naugle, Gregory; Sodhi, Manbir

    2014-01-01

    Multivariate statistical process monitoring (MSPM) is becoming increasingly utilized to further enhance process monitoring in the biopharmaceutical industry. MSPM can play a critical role when there are many measurements and these measurements are highly correlated, as is typical for many biopharmaceutical operations. Specifically, for processes such as cleaning-in-place (CIP) and steaming-in-place (SIP, also known as sterilization-in-place), control systems typically oversee the execution of the cycles, and verification of the outcome is based on offline assays. These offline assays add to delays and corrective actions may require additional setup times. Moreover, this conventional approach does not take interactive effects of process variables into account and cycle optimization opportunities as well as salient trends in the process may be missed. Therefore, more proactive and holistic online continued verification approaches are desirable. This article demonstrates the application of real-time MSPM to processes such as CIP and SIP with industrial examples. The proposed approach has significant potential for facilitating enhanced continuous verification, improved process understanding, abnormal situation detection, and predictive monitoring, as applied to CIP and SIP operations. © 2014 American Institute of Chemical Engineers.

  11. Measuring uncertainty in dose delivered to the cochlea due to setup error during external beam treatment of patients with cancer of the head and neck.

    PubMed

    Yan, M; Lovelock, D; Hunt, M; Mechalakos, J; Hu, Y; Pham, H; Jackson, A

    2013-12-01

    To use Cone Beam CT scans obtained just prior to treatments of head and neck cancer patients to measure the setup error and cumulative dose uncertainty of the cochlea. Data from 10 head and neck patients with 10 planning CTs and 52 Cone Beam CTs taken at time of treatment were used in this study. Patients were treated with conventional fractionation using an IMRT dose painting technique, most with 33 fractions. Weekly radiographic imaging was used to correct the patient setup. The authors used rigid registration of the planning CT and Cone Beam CT scans to find the translational and rotational setup errors, and the spatial setup errors of the cochlea. The planning CT was rotated and translated such that the cochlea positions match those seen in the cone beam scans, cochlea doses were recalculated and fractional doses accumulated. Uncertainties in the positions and cumulative doses of the cochlea were calculated with and without setup adjustments from radiographic imaging. The mean setup error of the cochlea was 0.04 ± 0.33 or 0.06 ± 0.43 cm for RL, 0.09 ± 0.27 or 0.07 ± 0.48 cm for AP, and 0.00 ± 0.21 or -0.24 ± 0.45 cm for SI with and without radiographic imaging, respectively. Setup with radiographic imaging reduced the standard deviation of the setup error by roughly 1-2 mm. The uncertainty of the cochlea dose depends on the treatment plan and the relative positions of the cochlea and target volumes. Combining results for the left and right cochlea, the authors found the accumulated uncertainty of the cochlea dose per fraction was 4.82 (0.39-16.8) cGy, or 10.1 (0.8-32.4) cGy, with and without radiographic imaging, respectively; the percentage uncertainties relative to the planned doses were 4.32% (0.28%-9.06%) and 10.2% (0.7%-63.6%), respectively. Patient setup error introduces uncertainty in the position of the cochlea during radiation treatment. With the assistance of radiographic imaging during setup, the standard deviation of setup error reduced by 31%, 42%, and 54% in RL, AP, and SI direction, respectively, and consequently, the uncertainty of the mean dose to cochlea reduced more than 50%. The authors estimate that the effects of these uncertainties on the probability of hearing loss for an individual patient could be as large as 10%.

  12. Measuring uncertainty in dose delivered to the cochlea due to setup error during external beam treatment of patients with cancer of the head and neck

    PubMed Central

    Yan, M.; Lovelock, D.; Hunt, M.; Mechalakos, J.; Hu, Y.; Pham, H.; Jackson, A.

    2013-01-01

    Purpose: To use Cone Beam CT scans obtained just prior to treatments of head and neck cancer patients to measure the setup error and cumulative dose uncertainty of the cochlea. Methods: Data from 10 head and neck patients with 10 planning CTs and 52 Cone Beam CTs taken at time of treatment were used in this study. Patients were treated with conventional fractionation using an IMRT dose painting technique, most with 33 fractions. Weekly radiographic imaging was used to correct the patient setup. The authors used rigid registration of the planning CT and Cone Beam CT scans to find the translational and rotational setup errors, and the spatial setup errors of the cochlea. The planning CT was rotated and translated such that the cochlea positions match those seen in the cone beam scans, cochlea doses were recalculated and fractional doses accumulated. Uncertainties in the positions and cumulative doses of the cochlea were calculated with and without setup adjustments from radiographic imaging. Results: The mean setup error of the cochlea was 0.04 ± 0.33 or 0.06 ± 0.43 cm for RL, 0.09 ± 0.27 or 0.07 ± 0.48 cm for AP, and 0.00 ± 0.21 or −0.24 ± 0.45 cm for SI with and without radiographic imaging, respectively. Setup with radiographic imaging reduced the standard deviation of the setup error by roughly 1–2 mm. The uncertainty of the cochlea dose depends on the treatment plan and the relative positions of the cochlea and target volumes. Combining results for the left and right cochlea, the authors found the accumulated uncertainty of the cochlea dose per fraction was 4.82 (0.39–16.8) cGy, or 10.1 (0.8–32.4) cGy, with and without radiographic imaging, respectively; the percentage uncertainties relative to the planned doses were 4.32% (0.28%–9.06%) and 10.2% (0.7%–63.6%), respectively. Conclusions: Patient setup error introduces uncertainty in the position of the cochlea during radiation treatment. With the assistance of radiographic imaging during setup, the standard deviation of setup error reduced by 31%, 42%, and 54% in RL, AP, and SI direction, respectively, and consequently, the uncertainty of the mean dose to cochlea reduced more than 50%. The authors estimate that the effects of these uncertainties on the probability of hearing loss for an individual patient could be as large as 10%. PMID:24320510

  13. PSpice Modeling of a Sandwich Piezoelectric Ceramic Ultrasonic Transducer in Longitudinal Vibration

    PubMed Central

    Wei, Xiaoyuan; Yang, Yuan; Yao, Wenqing; Zhang, Lei

    2017-01-01

    Sandwiched piezoelectric transducers are widely used, especially in high power applications. For more convenient analysis and design, a PSpice lossy model of sandwiched piezoelectric ultrasonic transducers in longitudinal vibration is proposed by means of the one-dimensional wave and transmission line theories. With the proposed model, the resonance and antiresonance frequencies are obtained, and it is shown that the simulations and measurements have good consistency. For the purpose of further verification the accuracy and application of the PSpice model, a pitch-catch setup and an experimental platform are built. They include two sandwiched piezoelectric ultrasonic transducers and two aluminum cylinders whose lengths are 20 mm and 100 mm respectively. Based on this pitch-catch setup, the impedance and transient analysis are performed. Compared with the measured results, it is shown that the simulated results have good consistency. In addition, the conclusion can be drawn that the optimal excitation frequency for the pitch-catch setup is not necessarily the resonance frequency of ultrasonic transducers, because the resonance frequency is obtained under no load. The proposed PSpice model of the sandwiched piezoelectric transducer is more conveniently applied to combine with other circuits such as driving circuits, filters, amplifiers, and so on. PMID:28973996

  14. PSpice Modeling of a Sandwich Piezoelectric Ceramic Ultrasonic Transducer in Longitudinal Vibration.

    PubMed

    Wei, Xiaoyuan; Yang, Yuan; Yao, Wenqing; Zhang, Lei

    2017-09-30

    Sandwiched piezoelectric transducers are widely used, especially in high power applications. For more convenient analysis and design, a PSpice lossy model of sandwiched piezoelectric ultrasonic transducers in longitudinal vibration is proposed by means of the one-dimensional wave and transmission line theories. With the proposed model, the resonance and antiresonance frequencies are obtained, and it is shown that the simulations and measurements have good consistency. For the purpose of further verification the accuracy and application of the PSpice model, a pitch-catch setup and an experimental platform are built. They include two sandwiched piezoelectric ultrasonic transducers and two aluminum cylinders whose lengths are 20 mm and 100 mm respectively. Based on this pitch-catch setup, the impedance and transient analysis are performed. Compared with the measured results, it is shown that the simulated results have good consistency. In addition, the conclusion can be drawn that the optimal excitation frequency for the pitch-catch setup is not necessarily the resonance frequency of ultrasonic transducers, because the resonance frequency is obtained under no load. The proposed PSpice model of the sandwiched piezoelectric transducer is more conveniently applied to combine with other circuits such as driving circuits, filters, amplifiers, and so on.

  15. Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Bert, Christoph; Metheany, Katherine G.; Doppke, Karen P.

    2006-03-15

    Purpose: To assess the utility of surface imaging on patient setup for accelerated partial-breast irradiation (APBI). Methods and Material: A photogrammetry system was used in parallel to APBI setup by laser and portal imaging. Surface data were acquired after laser and port-film setup for 9 patients. Surfaces were analyzed in comparison to a reference surface from the first treatment session by use of rigid transformations. The surface model after laser setup was used in a simulated photogrammetry setup procedure. In addition, breathing data were acquired by surface acquisition at a frame rate of 7 Hz. Results: Mean 3D displacement wasmore » 7.3 mm (SD, 4.4 mm) and 7.6 mm (SD, 4.2 mm) for laser and port film, respectively. Simulated setup with the photogrammetry system yielded mean displacement of 1 mm (SD, 1.2 mm). Distance analysis resulted in mean distances of 3.7 mm (SD, 4.9 mm), 4.3 mm (SD, 5.6 mm), and 1.6 mm (SD, 2.4 mm) for laser, port film, and photogrammetry, respectively. Breathing motion at isocenter was smaller than 3.7 mm, with a mean of 1.9 mm (SD, 1.1 mm). Conclusions: Surface imaging for PBI setup appears promising. Alignment of the 3D breast surface achieved by stereo-photogrammetry shows greater breast topology congruence than when patients are set up by laser or portal imaging. A correlation of breast surface and CTV must be quantitatively established.« less

  16. Comparison between infrared optical and stereoscopic X-ray technologies for patient setup in image guided stereotactic radiotherapy.

    PubMed

    Tagaste, Barbara; Riboldi, Marco; Spadea, Maria F; Bellante, Simone; Baroni, Guido; Cambria, Raffaella; Garibaldi, Cristina; Ciocca, Mario; Catalano, Gianpiero; Alterio, Daniela; Orecchia, Roberto

    2012-04-01

    To compare infrared (IR) optical vs. stereoscopic X-ray technologies for patient setup in image-guided stereotactic radiotherapy. Retrospective data analysis of 233 fractions in 127 patients treated with hypofractionated stereotactic radiotherapy was performed. Patient setup at the linear accelerator was carried out by means of combined IR optical localization and stereoscopic X-ray image fusion in 6 degrees of freedom (6D). Data were analyzed to evaluate the geometric and dosimetric discrepancy between the two patient setup strategies. Differences between IR optical localization and 6D X-ray image fusion parameters were on average within the expected localization accuracy, as limited by CT image resolution (3 mm). A disagreement between the two systems below 1 mm in all directions was measured in patients treated for cranial tumors. In extracranial sites, larger discrepancies and higher variability were observed as a function of the initial patient alignment. The compensation of IR-detected rotational errors resulted in a significantly improved agreement with 6D X-ray image fusion. On the basis of the bony anatomy registrations, the measured differences were found not to be sensitive to patient breathing. The related dosimetric analysis showed that IR-based patient setup caused limited variations in three cases, with 7% maximum dose reduction in the clinical target volume and no dose increase in organs at risk. In conclusion, patient setup driven by IR external surrogates localization in 6D featured comparable accuracy with respect to procedures based on stereoscopic X-ray imaging. Copyright © 2012 Elsevier Inc. All rights reserved.

  17. Automated patient setup and gating using cone beam computed tomography projections

    NASA Astrophysics Data System (ADS)

    Wan, Hanlin; Bertholet, Jenny; Ge, Jiajia; Poulsen, Per; Parikh, Parag

    2016-03-01

    In radiation therapy, fiducial markers are often implanted near tumors and used for patient positioning and respiratory gating purposes. These markers are then used to manually align the patients by matching the markers in the cone beam computed tomography (CBCT) reconstruction to those in the planning CT. This step is time-intensive and user-dependent, and often results in a suboptimal patient setup. We propose a fully automated, robust method based on dynamic programming (DP) for segmenting radiopaque fiducial markers in CBCT projection images, which are then used to automatically optimize the treatment couch position and/or gating window bounds. The mean of the absolute 2D segmentation error of our DP algorithm is 1.3+/- 1.0 mm for 87 markers on 39 patients. Intrafraction images were acquired every 3 s during treatment at two different institutions. For gated patients from Institution A (8 patients, 40 fractions), the DP algorithm increased the delivery accuracy (96+/- 6% versus 91+/- 11% , p  <  0.01) compared to the manual setup using kV fluoroscopy. For non-gated patients from Institution B (6 patients, 16 fractions), the DP algorithm performed similarly (1.5+/- 0.8 mm versus 1.6+/- 0.9 mm, p  =  0.48) compared to the manual setup matching the fiducial markers in the CBCT to the mean position. Our proposed automated patient setup algorithm only takes 1-2 s to run, requires no user intervention, and performs as well as or better than the current clinical setup.

  18. On-Line Use of Three-Dimensional Marker Trajectory Estimation From Cone-Beam Computed Tomography Projections for Precise Setup in Radiotherapy for Targets With Respiratory Motion

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Worm, Esben S., E-mail: esbeworm@rm.dk; Department of Medical Physics, Aarhus University Hospital, Aarhus; Hoyer, Morten

    2012-05-01

    Purpose: To develop and evaluate accurate and objective on-line patient setup based on a novel semiautomatic technique in which three-dimensional marker trajectories were estimated from two-dimensional cone-beam computed tomography (CBCT) projections. Methods and Materials: Seven treatment courses of stereotactic body radiotherapy for liver tumors were delivered in 21 fractions in total to 6 patients by a linear accelerator. Each patient had two to three gold markers implanted close to the tumors. Before treatment, a CBCT scan with approximately 675 two-dimensional projections was acquired during a full gantry rotation. The marker positions were segmented in each projection. From this, the three-dimensionalmore » marker trajectories were estimated using a probability based method. The required couch shifts for patient setup were calculated from the mean marker positions along the trajectories. A motion phantom moving with known tumor trajectories was used to examine the accuracy of the method. Trajectory-based setup was retrospectively used off-line for the first five treatment courses (15 fractions) and on-line for the last two treatment courses (6 fractions). Automatic marker segmentation was compared with manual segmentation. The trajectory-based setup was compared with setup based on conventional CBCT guidance on the markers (first 15 fractions). Results: Phantom measurements showed that trajectory-based estimation of the mean marker position was accurate within 0.3 mm. The on-line trajectory-based patient setup was performed within approximately 5 minutes. The automatic marker segmentation agreed with manual segmentation within 0.36 {+-} 0.50 pixels (mean {+-} SD; pixel size, 0.26 mm in isocenter). The accuracy of conventional volumetric CBCT guidance was compromised by motion smearing ({<=}21 mm) that induced an absolute three-dimensional setup error of 1.6 {+-} 0.9 mm (maximum, 3.2) relative to trajectory-based setup. Conclusions: The first on-line clinical use of trajectory estimation from CBCT projections for precise setup in stereotactic body radiotherapy was demonstrated. Uncertainty in the conventional CBCT-based setup procedure was eliminated with the new method.« less

  19. Time trend of injection drug errors before and after implementation of bar-code verification system.

    PubMed

    Sakushima, Ken; Umeki, Reona; Endoh, Akira; Ito, Yoichi M; Nasuhara, Yasuyuki

    2015-01-01

    Bar-code technology, used for verification of patients and their medication, could prevent medication errors in clinical practice. Retrospective analysis of electronically stored medical error reports was conducted in a university hospital. The number of reported medication errors of injected drugs, including wrong drug administration and administration to the wrong patient, was compared before and after implementation of the bar-code verification system for inpatient care. A total of 2867 error reports associated with injection drugs were extracted. Wrong patient errors decreased significantly after implementation of the bar-code verification system (17.4/year vs. 4.5/year, p< 0.05), although wrong drug errors did not decrease sufficiently (24.2/year vs. 20.3/year). The source of medication errors due to wrong drugs was drug preparation in hospital wards. Bar-code medication administration is effective for prevention of wrong patient errors. However, ordinary bar-code verification systems are limited in their ability to prevent incorrect drug preparation in hospital wards.

  20. Formal verification of medical monitoring software using Z language: a representative sample.

    PubMed

    Babamir, Seyed Morteza; Borhani, Mehdi

    2012-08-01

    Medical monitoring systems are useful aids assisting physicians in keeping patients under constant surveillance; however, taking sound decision by the systems is a physician concern. As a result, verification of the systems behavior in monitoring patients is a matter of significant. The patient monitoring is undertaken by software in modern medical systems; so, software verification of modern medial systems have been noticed. Such verification can be achieved by the Formal Languages having mathematical foundations. Among others, the Z language is a suitable formal language has been used to formal verification of systems. This study aims to present a constructive method to verify a representative sample of a medical system by which the system is visually specified and formally verified against patient constraints stated in Z Language. Exploiting our past experience in formal modeling Continuous Infusion Insulin Pump (CIIP), we think of the CIIP system as a representative sample of medical systems in proposing our present study. The system is responsible for monitoring diabetic's blood sugar.

  1. Experimental verification of ‘waveguide’ plasmonics

    NASA Astrophysics Data System (ADS)

    Prudêncio, Filipa R.; Costa, Jorge R.; Fernandes, Carlos A.; Engheta, Nader; Silveirinha, Mário G.

    2017-12-01

    Surface plasmons polaritons are collective excitations of an electron gas that occur at an interface between negative-ɛ and positive-ɛ media. Here, we report the experimental observation of such surface waves using simple waveguide metamaterials filled only with available positive-ɛ media at microwave frequencies. In contrast to optical designs, in our setup the propagation length of the surface plasmons can be rather long as low loss conventional dielectrics are chosen to avoid typical losses from negative-ɛ media. Plasmonic phenomena have potential applications in enhancing light-matter interactions, implementing nanoscale photonic circuits and integrated photonics.

  2. In vivo dosimetry in external beam radiotherapy.

    PubMed

    Mijnheer, Ben; Beddar, Sam; Izewska, Joanna; Reft, Chester

    2013-07-01

    In vivo dosimetry (IVD) is in use in external beam radiotherapy (EBRT) to detect major errors, to assess clinically relevant differences between planned and delivered dose, to record dose received by individual patients, and to fulfill legal requirements. After discussing briefly the main characteristics of the most commonly applied IVD systems, the clinical experience of IVD during EBRT will be summarized. Advancement of the traditional aspects of in vivo dosimetry as well as the development of currently available and newly emerging noninterventional technologies are required for large-scale implementation of IVD in EBRT. These new technologies include the development of electronic portal imaging devices for 2D and 3D patient dosimetry during advanced treatment techniques, such as IMRT and VMAT, and the use of IVD in proton and ion radiotherapy by measuring the decay of radiation-induced radionuclides. In the final analysis, we will show in this Vision 20∕20 paper that in addition to regulatory compliance and reimbursement issues, the rationale for in vivo measurements is to provide an accurate and independent verification of the overall treatment procedure. It will enable the identification of potential errors in dose calculation, data transfer, dose delivery, patient setup, and changes in patient anatomy. It is the authors' opinion that all treatments with curative intent should be verified through in vivo dose measurements in combination with pretreatment checks.

  3. Technical Note: Range verification system using edge detection method for a scintillator and a CCD camera system.

    PubMed

    Saotome, Naoya; Furukawa, Takuji; Hara, Yousuke; Mizushima, Kota; Tansho, Ryohei; Saraya, Yuichi; Shirai, Toshiyuki; Noda, Koji

    2016-04-01

    Three-dimensional irradiation with a scanned carbon-ion beam has been performed from 2011 at the authors' facility. The authors have developed the rotating-gantry equipped with the scanning irradiation system. The number of combinations of beam properties to measure for the commissioning is more than 7200, i.e., 201 energy steps, 3 intensities, and 12 gantry angles. To compress the commissioning time, quick and simple range verification system is required. In this work, the authors develop a quick range verification system using scintillator and charge-coupled device (CCD) camera and estimate the accuracy of the range verification. A cylindrical plastic scintillator block and a CCD camera were installed on the black box. The optical spatial resolution of the system is 0.2 mm/pixel. The camera control system was connected and communicates with the measurement system that is part of the scanning system. The range was determined by image processing. Reference range for each energy beam was determined by a difference of Gaussian (DOG) method and the 80% of distal dose of the depth-dose distribution that were measured by a large parallel-plate ionization chamber. The authors compared a threshold method and a DOG method. The authors found that the edge detection method (i.e., the DOG method) is best for the range detection. The accuracy of range detection using this system is within 0.2 mm, and the reproducibility of the same energy measurement is within 0.1 mm without setup error. The results of this study demonstrate that the authors' range check system is capable of quick and easy range verification with sufficient accuracy.

  4. Technical Note: Range verification system using edge detection method for a scintillator and a CCD camera system

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Saotome, Naoya, E-mail: naosao@nirs.go.jp; Furukawa, Takuji; Hara, Yousuke

    Purpose: Three-dimensional irradiation with a scanned carbon-ion beam has been performed from 2011 at the authors’ facility. The authors have developed the rotating-gantry equipped with the scanning irradiation system. The number of combinations of beam properties to measure for the commissioning is more than 7200, i.e., 201 energy steps, 3 intensities, and 12 gantry angles. To compress the commissioning time, quick and simple range verification system is required. In this work, the authors develop a quick range verification system using scintillator and charge-coupled device (CCD) camera and estimate the accuracy of the range verification. Methods: A cylindrical plastic scintillator blockmore » and a CCD camera were installed on the black box. The optical spatial resolution of the system is 0.2 mm/pixel. The camera control system was connected and communicates with the measurement system that is part of the scanning system. The range was determined by image processing. Reference range for each energy beam was determined by a difference of Gaussian (DOG) method and the 80% of distal dose of the depth-dose distribution that were measured by a large parallel-plate ionization chamber. The authors compared a threshold method and a DOG method. Results: The authors found that the edge detection method (i.e., the DOG method) is best for the range detection. The accuracy of range detection using this system is within 0.2 mm, and the reproducibility of the same energy measurement is within 0.1 mm without setup error. Conclusions: The results of this study demonstrate that the authors’ range check system is capable of quick and easy range verification with sufficient accuracy.« less

  5. Couch height–based patient setup for abdominal radiation therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ohira, Shingo; Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita; Ueda, Yoshihiro

    2016-04-01

    There are 2 methods commonly used for patient positioning in the anterior-posterior (A-P) direction: one is the skin mark patient setup method (SMPS) and the other is the couch height–based patient setup method (CHPS). This study compared the setup accuracy of these 2 methods for abdominal radiation therapy. The enrollment for this study comprised 23 patients with pancreatic cancer. For treatments (539 sessions), patients were set up by using isocenter skin marks and thereafter treatment couch was shifted so that the distance between the isocenter and the upper side of the treatment couch was equal to that indicated on themore » computed tomographic (CT) image. Setup deviation in the A-P direction for CHPS was measured by matching the spine of the digitally reconstructed radiograph (DRR) of a lateral beam at simulation with that of the corresponding time-integrated electronic portal image. For SMPS with no correction (SMPS/NC), setup deviation was calculated based on the couch-level difference between SMPS and CHPS. SMPS/NC was corrected using 2 off-line correction protocols: no action level (SMPS/NAL) and extended NAL (SMPS/eNAL) protocols. Margins to compensate for deviations were calculated using the Stroom formula. A-P deviation > 5 mm was observed in 17% of SMPS/NC, 4% of SMPS/NAL, and 4% of SMPS/eNAL sessions but only in one CHPS session. For SMPS/NC, 7 patients (30%) showed deviations at an increasing rate of > 0.1 mm/fraction, but for CHPS, no such trend was observed. The standard deviations (SDs) of systematic error (Σ) were 2.6, 1.4, 0.6, and 0.8 mm and the root mean squares of random error (σ) were 2.1, 2.6, 2.7, and 0.9 mm for SMPS/NC, SMPS/NAL, SMPS/eNAL, and CHPS, respectively. Margins to compensate for the deviations were wide for SMPS/NC (6.7 mm), smaller for SMPS/NAL (4.6 mm) and SMPS/eNAL (3.1 mm), and smallest for CHPS (2.2 mm). Achieving better setup with smaller margins, CHPS appears to be a reproducible method for abdominal patient setup.« less

  6. Analysis of Prostate Patient Setup and Tracking Data: Potential Intervention Strategies

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Su Zhong, E-mail: zsu@floridaproton.org; Zhang Lisha; Murphy, Martin

    Purpose: To evaluate the setup, interfraction, and intrafraction organ motion error distributions and simulate intrafraction intervention strategies for prostate radiotherapy. Methods and Materials: A total of 17 patients underwent treatment setup and were monitored using the Calypso system during radiotherapy. On average, the prostate tracking measurements were performed for 8 min/fraction for 28 fractions for each patient. For both patient couch shift data and intrafraction organ motion data, the systematic and random errors were obtained from the patient population. The planning target volume margins were calculated using the van Herk formula. Two intervention strategies were simulated using the tracking data:more » the deviation threshold and period. The related planning target volume margins, time costs, and prostate position 'fluctuation' were presented. Results: The required treatment margin for the left-right, superoinferior, and anteroposterior axes was 8.4, 10.8, and 14.7 mm for skin mark-only setup and 1.3, 2.3, and 2.8 mm using the on-line setup correction, respectively. Prostate motion significantly correlated among the superoinferior and anteroposterior directions. Of the 17 patients, 14 had prostate motion within 5 mm of the initial setup position for {>=}91.6% of the total tracking time. The treatment margin decreased to 1.1, 1.8, and 2.3 mm with a 3-mm threshold correction and to 0.5, 1.0, and 1.5 mm with an every-2-min correction in the left-right, superoinferior, and anteroposterior directions, respectively. The periodic corrections significantly increase the treatment time and increased the number of instances when the setup correction was made during transient excursions. Conclusions: The residual systematic and random error due to intrafraction prostate motion is small after on-line setup correction. Threshold-based and time-based intervention strategies both reduced the planning target volume margins. The time-based strategies increased the treatment time and the in-fraction position fluctuation.« less

  7. TU-FG-201-03: Automatic Pre-Delivery Verification Using Statistical Analysis of Consistencies in Treatment Plan Parameters by the Treatment Site and Modality

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Liu, S; Wu, Y; Chang, X

    Purpose: A novel computer software system, namely APDV (Automatic Pre-Delivery Verification), has been developed for verifying patient treatment plan parameters right prior to treatment deliveries in order to automatically detect and prevent catastrophic errors. Methods: APDV is designed to continuously monitor new DICOM plan files on the TMS computer at the treatment console. When new plans to be delivered are detected, APDV checks the consistencies of plan parameters and high-level plan statistics using underlying rules and statistical properties based on given treatment site, technique and modality. These rules were quantitatively derived by retrospectively analyzing all the EBRT treatment plans ofmore » the past 8 years at authors’ institution. Therapists and physicists will be notified with a warning message displayed on the TMS computer if any critical errors are detected, and check results, confirmation, together with dismissal actions will be saved into database for further review. Results: APDV was implemented as a stand-alone program using C# to ensure required real time performance. Mean values and standard deviations were quantitatively derived for various plan parameters including MLC usage, MU/cGy radio, beam SSD, beam weighting, and the beam gantry angles (only for lateral targets) per treatment site, technique and modality. 2D-based rules of combined MU/cGy ratio and averaged SSD values were also derived using joint probabilities of confidence error ellipses. The statistics of these major treatment plan parameters quantitatively evaluate the consistency of any treatment plans which facilitates automatic APDV checking procedures. Conclusion: APDV could be useful in detecting and preventing catastrophic errors immediately before treatment deliveries. Future plan including automatic patient identify and patient setup checks after patient daily images are acquired by the machine and become available on the TMS computer. This project is supported by the Agency for Healthcare Research and Quality (AHRQ) under award 1R01HS0222888. The senior author received research grants from ViewRay Inc. and Varian Medical System.« less

  8. Positioning accuracy for lung stereotactic body radiotherapy patients determined by on-treatment cone-beam CT imaging

    PubMed Central

    Richmond, N D; Pilling, K E; Peedell, C; Shakespeare, D; Walker, C P

    2012-01-01

    Stereotactic body radiotherapy for early stage non-small cell lung cancer is an emerging treatment option in the UK. Since relatively few high-dose ablative fractions are delivered to a small target volume, the consequences of a geometric miss are potentially severe. This paper presents the results of treatment delivery set-up data collected using Elekta Synergy (Elekta, Crawley, UK) cone-beam CT imaging for 17 patients immobilised using the Bodyfix system (Medical Intelligence, Schwabmuenchen, Germany). Images were acquired on the linear accelerator at initial patient treatment set-up, following any position correction adjustments, and post-treatment. These were matched to the localisation CT scan using the Elekta XVI software. In total, 71 fractions were analysed for patient set-up errors. The mean vector error at initial set-up was calculated as 5.3±2.7 mm, which was significantly reduced to 1.4±0.7 mm following image guided correction. Post-treatment the corresponding value was 2.1±1.2 mm. The use of the Bodyfix abdominal compression plate on 5 patients to reduce the range of tumour excursion during respiration produced mean longitudinal set-up corrections of −4.4±4.5 mm compared with −0.7±2.6 mm without compression for the remaining 12 patients. The use of abdominal compression led to a greater variation in set-up errors and a shift in the mean value. PMID:22665927

  9. Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Siebers, Jeffrey V.; Keall, Paul J.; Wu Qiuwen

    2005-10-01

    Purpose: The purpose of this study is to determine dose delivery errors that could result from random and systematic setup errors for head-and-neck patients treated using the simultaneous integrated boost (SIB)-intensity-modulated radiation therapy (IMRT) technique. Methods and Materials: Twenty-four patients who participated in an intramural Phase I/II parotid-sparing IMRT dose-escalation protocol using the SIB treatment technique had their dose distributions reevaluated to assess the impact of random and systematic setup errors. The dosimetric effect of random setup error was simulated by convolving the two-dimensional fluence distribution of each beam with the random setup error probability density distribution. Random setup errorsmore » of {sigma} = 1, 3, and 5 mm were simulated. Systematic setup errors were simulated by randomly shifting the patient isocenter along each of the three Cartesian axes, with each shift selected from a normal distribution. Systematic setup error distributions with {sigma} = 1.5 and 3.0 mm along each axis were simulated. Combined systematic and random setup errors were simulated for {sigma} = {sigma} = 1.5 and 3.0 mm along each axis. For each dose calculation, the gross tumor volume (GTV) received by 98% of the volume (D{sub 98}), clinical target volume (CTV) D{sub 90}, nodes D{sub 90}, cord D{sub 2}, and parotid D{sub 50} and parotid mean dose were evaluated with respect to the plan used for treatment for the structure dose and for an effective planning target volume (PTV) with a 3-mm margin. Results: Simultaneous integrated boost-IMRT head-and-neck treatment plans were found to be less sensitive to random setup errors than to systematic setup errors. For random-only errors, errors exceeded 3% only when the random setup error {sigma} exceeded 3 mm. Simulated systematic setup errors with {sigma} = 1.5 mm resulted in approximately 10% of plan having more than a 3% dose error, whereas a {sigma} = 3.0 mm resulted in half of the plans having more than a 3% dose error and 28% with a 5% dose error. Combined random and systematic dose errors with {sigma} = {sigma} = 3.0 mm resulted in more than 50% of plans having at least a 3% dose error and 38% of the plans having at least a 5% dose error. Evaluation with respect to a 3-mm expanded PTV reduced the observed dose deviations greater than 5% for the {sigma} = {sigma} = 3.0 mm simulations to 5.4% of the plans simulated. Conclusions: Head-and-neck SIB-IMRT dosimetric accuracy would benefit from methods to reduce patient systematic setup errors. When GTV, CTV, or nodal volumes are used for dose evaluation, plans simulated including the effects of random and systematic errors deviate substantially from the nominal plan. The use of PTVs for dose evaluation in the nominal plan improves agreement with evaluated GTV, CTV, and nodal dose values under simulated setup errors. PTV concepts should be used for SIB-IMRT head-and-neck squamous cell carcinoma patients, although the size of the margins may be less than those used with three-dimensional conformal radiation therapy.« less

  10. Suite of Benchmark Tests to Conduct Mesh-Convergence Analysis of Nonlinear and Non-constant Coefficient Transport Codes

    NASA Astrophysics Data System (ADS)

    Zamani, K.; Bombardelli, F. A.

    2014-12-01

    Verification of geophysics codes is imperative to avoid serious academic as well as practical consequences. In case that access to any given source code is not possible, the Method of Manufactured Solution (MMS) cannot be employed in code verification. In contrast, employing the Method of Exact Solution (MES) has several practical advantages. In this research, we first provide four new one-dimensional analytical solutions designed for code verification; these solutions are able to uncover the particular imperfections of the Advection-diffusion-reaction equation, such as nonlinear advection, diffusion or source terms, as well as non-constant coefficient equations. After that, we provide a solution of Burgers' equation in a novel setup. Proposed solutions satisfy the continuity of mass for the ambient flow, which is a crucial factor for coupled hydrodynamics-transport solvers. Then, we use the derived analytical solutions for code verification. To clarify gray-literature issues in the verification of transport codes, we designed a comprehensive test suite to uncover any imperfection in transport solvers via a hierarchical increase in the level of tests' complexity. The test suite includes hundreds of unit tests and system tests to check vis-a-vis the portions of the code. Examples for checking the suite start by testing a simple case of unidirectional advection; then, bidirectional advection and tidal flow and build up to nonlinear cases. We design tests to check nonlinearity in velocity, dispersivity and reactions. The concealing effect of scales (Peclet and Damkohler numbers) on the mesh-convergence study and appropriate remedies are also discussed. For the cases in which the appropriate benchmarks for mesh convergence study are not available, we utilize symmetry. Auxiliary subroutines for automation of the test suite and report generation are designed. All in all, the test package is not only a robust tool for code verification but it also provides comprehensive insight on the ADR solvers capabilities. Such information is essential for any rigorous computational modeling of ADR equation for surface/subsurface pollution transport. We also convey our experiences in finding several errors which were not detectable with routine verification techniques.

  11. Dosimetry in radiotherapy using a-Si EPIDs: Systems, methods, and applications focusing on 3D patient dose estimation

    NASA Astrophysics Data System (ADS)

    McCurdy, B. M. C.

    2013-06-01

    An overview is provided of the use of amorphous silicon electronic portal imaging devices (EPIDs) for dosimetric purposes in radiation therapy, focusing on 3D patient dose estimation. EPIDs were originally developed to provide on-treatment radiological imaging to assist with patient setup, but there has also been a natural interest in using them as dosimeters since they use the megavoltage therapy beam to form images. The current generation of clinically available EPID technology, amorphous-silicon (a-Si) flat panel imagers, possess many characteristics that make them much better suited to dosimetric applications than earlier EPID technologies. Features such as linearity with dose/dose rate, high spatial resolution, realtime capability, minimal optical glare, and digital operation combine with the convenience of a compact, retractable detector system directly mounted on the linear accelerator to provide a system that is well-suited to dosimetric applications. This review will discuss clinically available a-Si EPID systems, highlighting dosimetric characteristics and remaining limitations. Methods for using EPIDs in dosimetry applications will be discussed. Dosimetric applications using a-Si EPIDs to estimate three-dimensional dose in the patient during treatment will be overviewed. Clinics throughout the world are implementing increasingly complex treatments such as dynamic intensity modulated radiation therapy and volumetric modulated arc therapy, as well as specialized treatment techniques using large doses per fraction and short treatment courses (ie. hypofractionation and stereotactic radiosurgery). These factors drive the continued strong interest in using EPIDs as dosimeters for patient treatment verification.

  12. The effect of systematic set-up deviations on the absorbed dose distribution for left-sided breast cancer treated with respiratory gating

    NASA Astrophysics Data System (ADS)

    Edvardsson, A.; Ceberg, S.

    2013-06-01

    The aim of this study was 1) to investigate interfraction set-up uncertainties for patients treated with respiratory gating for left-sided breast cancer, 2) to investigate the effect of the inter-fraction set-up on the absorbed dose-distribution for the target and organs at risk (OARs) and 3) optimize the set-up correction strategy. By acquiring multiple set-up images the systematic set-up deviation was evaluated. The effect of the systematic set-up deviation on the absorbed dose distribution was evaluated by 1) simulation in the treatment planning system and 2) measurements with a biplanar diode array. The set-up deviations could be decreased using a no action level correction strategy. Not using the clinically implemented adaptive maximum likelihood factor for the gating patients resulted in better set-up. When the uncorrected set-up deviations were simulated the average mean absorbed dose was increased from 1.38 to 2.21 Gy for the heart, 4.17 to 8.86 Gy to the left anterior descending coronary artery and 5.80 to 7.64 Gy to the left lung. Respiratory gating can induce systematic set-up deviations which would result in increased mean absorbed dose to the OARs if not corrected for and should therefore be corrected for by an appropriate correction strategy.

  13. SU-E-J-21: Setup Variability of Colorectal Cancer Patients Treated in the Prone Position and Dosimetric Comparison with the Supine Position

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kim, A; Foster, J; Chu, W

    2015-06-15

    Purpose: Many cancer centers treat colorectal patients in the prone position on a belly board to minimize dose to the small bowel. That may potentially Result in patient setup instability with corresponding impact on dose delivery accuracy for highly conformal techniques such as IMRT/VMAT. Two aims of this work are 1) to investigate setup accuracy of rectum patients treated in the prone position on a belly board using CBCT and 2) to evaluate dosimetric impact on bladder and small bowel of treating rectum patients in supine vs. prone position. Methods: For the setup accuracy study, 10 patients were selected. Weeklymore » CBCTs were acquired and matched to bone. The CBCT-determined shifts were recorded. For the dosimetric study, 7 prone-setup patients and 7 supine-setup patients were randomly selected from our clinical database. Various clinically relevant dose volume histogram values were recorded for the small bowel and bladder. Results: The CBCT-determined rotational shifts had a wide variation. For the dataset acquired at the time of this writing, the ranges of rotational setup errors for pitch, roll, and yaw were [−3.6° 4.7°], [−4.3° 3.2°], and [−1.4° 1.4°]. For the dosimetric study: the small bowel V(45Gy) and mean dose for the prone position was 5.6±12.1% and 18.4±6.2Gy (ranges indicate standard deviations); for the supine position the corresponding dose values were 12.9±15.8% and 24.7±8.8Gy. For the bladder, the V(30Gy) and mean dose for prone position were 68.7±12.7% and 38.4±3.3Gy; for supine position these dose values were 77.1±13.7% and 40.7±3.1Gy. Conclusion: There is evidence of significant rotational instability in the prone position. The OAR dosimetry study indicates that there are some patients that may still benefit from the prone position, though many patients can be safely treated supine.« less

  14. SU-F-J-142: Proposed Method to Broaden Inclusion Potential of Patients Able to Use the Calypso Tracking System in Prostate Radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fiedler, D; Kuo, H; Bodner, W

    2016-06-15

    Purpose: To introduce a non-standard method of patient setup, using BellyBoard immobilization, to better utilize the localization and tracking potential of an RF-beacon system with EBRT for prostate cancer. Methods: An RF-beacon phantom was imaged using a wide bore CT scanner, both in a standard level position and with a known rotation (4° pitch and 7.5° yaw). A commercial treatment planning system (TPS) was used to determine positional coordinates of each beacon, and the centroid of the three beacons for both setups. For each setup at the Linac, kV AP and Rt Lateral images were obtained. A full characterization ofmore » the RF-beacon system in clinical mode was completed for various beacons’ array-to-centroid distances, which includes vertical, lateral, and longitudinal offset data, as well as pitch and yaw offset measurements for the tilted phantom. For the single patient who has been setup using the proposed BellyBoard method, a supine simulation was first obtained. When abdominal protrusion was found to be exceeding the limits of the RF-Beacon system through distance-based analysis in the TPS, the patient is re-simulated prone with the BellyBoard. Array to centroid distance is measured again in the TPS, and if found to be within the localization or tracking region it is applied. Results: Characterization of limitations for the RF-beacon system in clinical mode showed acceptable consistency of offset determination for phantom setup accuracy. The nonstandard patient setup method reduced the beacons’ centroid-to-array distance by 8.32cm, from 25.13cm to 16.81cm; completely out of tracking range (greater than 20cm) to within setup tracking range (less than 20cm). Conclusion: Using the RF-beacon system in combination with this novel patient setup can allow patients who would otherwise not be candidates for beacon enhanced EBRT to now be able to benefit from the reduced PTV margins of this treatment method.« less

  15. Improvements of low-level radioxenon detection sensitivity by a state-of-the art coincidence setup.

    PubMed

    Cagniant, A; Le Petit, G; Gross, P; Douysset, G; Richard-Bressand, H; Fontaine, J-P

    2014-05-01

    The ability to quantify isotopic ratios of 135, 133 m, 133 and 131 m radioxenon is essential for the verification of the Comprehensive Nuclear-Test Ban Treaty (CTBT). In order to improve detection limits, CEA has developed a new on-site setup using photon/electron coincidence (Le Petit et al., 2013. J. Radioanal. Nucl. Chem., DOI : 10.1007/s 10697-013-2525-8.). Alternatively, the electron detection cell equipped with large silicon chips (PIPS) can be used with HPGe detector for laboratory analysis purpose. This setup allows the measurement of β/γ coincidences for the detection of (133)Xe and (135)Xe; and K-shell Conversion Electrons (K-CE)/X-ray coincidences for the detection of (131m)Xe, (133m)Xe and (133)Xe as well. Good energy resolution of 11 keV at 130 keV and low energy threshold of 29 keV for the electron detection were obtained. This provides direct discrimination between K-CE from (133)Xe, (133m)Xe and (131m)Xe. Estimation of Minimum Detectable Activity (MDA) for (131m)Xe is in the order of 1mBq over a 4 day measurement. An analysis of an environmental radioxenon sample using this method is shown. © 2013 The Authors. Published by Elsevier Ltd All rights reserved.

  16. Experimental verification of long-term evolution radio transmissions over dual-polarization combined fiber and free-space optics optical infrastructures.

    PubMed

    Bohata, J; Zvanovec, S; Pesek, P; Korinek, T; Mansour Abadi, M; Ghassemlooy, Z

    2016-03-10

    This paper describes the experimental verification of the utilization of long-term evolution radio over fiber (RoF) and radio over free space optics (RoFSO) systems using dual-polarization signals for cloud radio access network applications determining the specific utilization limits. A number of free space optics configurations are proposed and investigated under different atmospheric turbulence regimes in order to recommend the best setup configuration. We show that the performance of the proposed link, based on the combination of RoF and RoFSO for 64 QAM at 2.6 GHz, is more affected by the turbulence based on the measured difference error vector magnitude value of 5.5%. It is further demonstrated the proposed systems can offer higher noise immunity under particular scenarios with the signal-to-noise ratio reliability limit of 5 dB in the radio frequency domain for RoF and 19.3 dB in the optical domain for a combination of RoF and RoFSO links.

  17. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zhang, JY; Hong, DL

    Purpose: The purpose of this study is to investigate the patient set-up error and interfraction target coverage in cervical cancer using image-guided adaptive radiotherapy (IGART) with cone-beam computed tomography (CBCT). Methods: Twenty cervical cancer patients undergoing intensity modulated radiotherapy (IMRT) were randomly selected. All patients were matched to the isocenter using laser with the skin markers. Three dimensional CBCT projections were acquired by the Varian Truebeam treatment system. Set-up errors were evaluated by radiation oncologists, after CBCT correction. The clinical target volume (CTV) was delineated on each CBCT, and the planning target volume (PTV) coverage of each CBCT-CTVs was analyzed.more » Results: A total of 152 CBCT scans were acquired from twenty cervical cancer patients, the mean set-up errors in the longitudinal, vertical, and lateral direction were 3.57, 2.74 and 2.5mm respectively, without CBCT corrections. After corrections, these were decreased to 1.83, 1.44 and 0.97mm. For the target coverage, CBCT-CTV coverage without CBCT correction was 94% (143/152), and 98% (149/152) with correction. Conclusion: Use of CBCT verfication to measure patient setup errors could be applied to improve the treatment accuracy. In addition, the set-up error corrections significantly improve the CTV coverage for cervical cancer patients.« less

  18. Improved cyberinfrastructure for integrated hydrometeorological predictions within the fully-coupled WRF-Hydro modeling system

    NASA Astrophysics Data System (ADS)

    gochis, David; hooper, Rick; parodi, Antonio; Jha, Shantenu; Yu, Wei; Zaslavsky, Ilya; Ganapati, Dinesh

    2014-05-01

    The community WRF-Hydro system is currently being used in a variety of flood prediction and regional hydroclimate impacts assessment applications around the world. Despite its increasingly wide use certain cyberinfrastructure bottlenecks exist in the setup, execution and post-processing of WRF-Hydro model runs. These bottlenecks result in wasted time, labor, data transfer bandwidth and computational resource use. Appropriate development and use of cyberinfrastructure to setup and manage WRF-Hydro modeling applications will streamline the entire workflow of hydrologic model predictions. This talk will present recent advances in the development and use of new open-source cyberinfrastructure tools for the WRF-Hydro architecture. These tools include new web-accessible pre-processing applications, supercomputer job management applications and automated verification and visualization applications. The tools will be described successively and then demonstrated in a set of flash flood use cases for recent destructive flood events in the U.S. and in Europe. Throughout, an emphasis on the implementation and use of community data standards for data exchange is made.

  19. SU-F-P-18: Development of the Technical Training System for Patient Set-Up Considering Rotational Correction in the Virtual Environment Using Three-Dimensional Computer Graphic Engine

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Imura, K; Fujibuchi, T; Hirata, H

    Purpose: Patient set-up skills in radiotherapy treatment room have a great influence on treatment effect for image guided radiotherapy. In this study, we have developed the training system for improving practical set-up skills considering rotational correction in the virtual environment away from the pressure of actual treatment room by using three-dimensional computer graphic (3DCG) engine. Methods: The treatment room for external beam radiotherapy was reproduced in the virtual environment by using 3DCG engine (Unity). The viewpoints to perform patient set-up in the virtual treatment room were arranged in both sides of the virtual operable treatment couch to assume actual performancemore » by two clinical staffs. The position errors to mechanical isocenter considering alignment between skin marker and laser on the virtual patient model were displayed by utilizing numerical values expressed in SI units and the directions of arrow marks. The rotational errors calculated with a point on the virtual body axis as the center of each rotation axis for the virtual environment were corrected by adjusting rotational position of the body phantom wound the belt with gyroscope preparing on table in a real space. These rotational errors were evaluated by describing vector outer product operations and trigonometric functions in the script for patient set-up technique. Results: The viewpoints in the virtual environment allowed individual user to visually recognize the position discrepancy to mechanical isocenter until eliminating the positional errors of several millimeters. The rotational errors between the two points calculated with the center point could be efficiently corrected to display the minimum technique mathematically by utilizing the script. Conclusion: By utilizing the script to correct the rotational errors as well as accurate positional recognition for patient set-up technique, the training system developed for improving patient set-up skills enabled individual user to indicate efficient positional correction methods easily.« less

  20. Characterization of the microbunch time structure of proton pencil beams at a clinical treatment facility.

    PubMed

    Petzoldt, J; Roemer, K E; Enghardt, W; Fiedler, F; Golnik, C; Hueso-González, F; Helmbrecht, S; Kormoll, T; Rohling, H; Smeets, J; Werner, T; Pausch, G

    2016-03-21

    Proton therapy is an advantageous treatment modality compared to conventional radiotherapy. In contrast to photons, charged particles have a finite range and can thus spare organs at risk. Additionally, the increased ionization density in the so-called Bragg peak close to the particle range can be utilized for maximum dose deposition in the tumour volume. Unfortunately, the accuracy of the therapy can be affected by range uncertainties, which have to be covered by additional safety margins around the treatment volume. A real-time range and dose verification is therefore highly desired and would be key to exploit the major advantages of proton therapy. Prompt gamma rays, produced in nuclear reactions between projectile and target nuclei, can be used to measure the proton's range. The prompt gamma-ray timing (PGT) method aims at obtaining this information by determining the gamma-ray emission time along the proton path using a conventional time-of-flight detector setup. First tests at a clinical accelerator have shown the feasibility to observe range shifts of about 5 mm at clinically relevant doses. However, PGT spectra are smeared out by the bunch time spread. Additionally, accelerator related proton bunch drifts against the radio frequency have been detected, preventing a potential range verification. At OncoRay, first experiments using a proton bunch monitor (PBM) at a clinical pencil beam have been conducted. Elastic proton scattering at a hydrogen-containing foil could be utilized to create a coincident proton-proton signal in two identical PBMs. The selection of coincident events helped to suppress uncorrelated background. The PBM setup was used as time reference for a PGT detector to correct for potential bunch drifts. Furthermore, the corrected PGT data were used to image an inhomogeneous phantom. In a further systematic measurement campaign, the bunch time spread and the proton transmission rate were measured for several beam energies between 69 and 225 MeV as well as for variable momentum limiting slit openings. We conclude that the usage of a PBM increases the robustness of the PGT method in clinical conditions and that the obtained data will help to create reliable range verification procedures in clinical routine.

  1. Comparison of measurement methods with a mixed effects procedure accounting for replicated evaluations (COM3PARE): method comparison algorithm implementation for head and neck IGRT positional verification.

    PubMed

    Roy, Anuradha; Fuller, Clifton D; Rosenthal, David I; Thomas, Charles R

    2015-08-28

    Comparison of imaging measurement devices in the absence of a gold-standard comparator remains a vexing problem; especially in scenarios where multiple, non-paired, replicated measurements occur, as in image-guided radiotherapy (IGRT). As the number of commercially available IGRT presents a challenge to determine whether different IGRT methods may be used interchangeably, an unmet need conceptually parsimonious and statistically robust method to evaluate the agreement between two methods with replicated observations. Consequently, we sought to determine, using an previously reported head and neck positional verification dataset, the feasibility and utility of a Comparison of Measurement Methods with the Mixed Effects Procedure Accounting for Replicated Evaluations (COM3PARE), a unified conceptual schema and analytic algorithm based upon Roy's linear mixed effects (LME) model with Kronecker product covariance structure in a doubly multivariate set-up, for IGRT method comparison. An anonymized dataset consisting of 100 paired coordinate (X/ measurements from a sequential series of head and neck cancer patients imaged near-simultaneously with cone beam CT (CBCT) and kilovoltage X-ray (KVX) imaging was used for model implementation. Software-suggested CBCT and KVX shifts for the lateral (X), vertical (Y) and longitudinal (Z) dimensions were evaluated for bias, inter-method (between-subject variation), intra-method (within-subject variation), and overall agreement using with a script implementing COM3PARE with the MIXED procedure of the statistical software package SAS (SAS Institute, Cary, NC, USA). COM3PARE showed statistically significant bias agreement and difference in inter-method between CBCT and KVX was observed in the Z-axis (both p - value<0.01). Intra-method and overall agreement differences were noted as statistically significant for both the X- and Z-axes (all p - value<0.01). Using pre-specified criteria, based on intra-method agreement, CBCT was deemed preferable for X-axis positional verification, with KVX preferred for superoinferior alignment. The COM3PARE methodology was validated as feasible and useful in this pilot head and neck cancer positional verification dataset. COM3PARE represents a flexible and robust standardized analytic methodology for IGRT comparison. The implemented SAS script is included to encourage other groups to implement COM3PARE in other anatomic sites or IGRT platforms.

  2. Design and implementation of a head-and-neck phantom for system audit and verification of intensity-modulated radiation therapy.

    PubMed

    Webster, Gareth J; Hardy, Mark J; Rowbottom, Carl G; Mackay, Ranald I

    2008-04-16

    The head and neck is a challenging anatomic site for intensity-modulated radiation therapy (IMRT), requiring thorough testing of planning and treatment delivery systems. Ideally, the phantoms used should be anatomically realistic, have radiologic properties identical to those of the tissues concerned, and allow for the use of a variety of devices to verify dose and dose distribution in any target or normaltissue structure. A phantom that approaches the foregoing characteristics has been designed and built; its specific purpose is verification for IMRT treatments in the head-andneck region. This semi-anatomic phantom, HANK, is constructed of Perspex (Imperial Chemical Industries, London, U.K.) and provides for the insertion of heterogeneities simulating air cavities in a range of fixed positions. Chamber inserts are manufactured to incorporate either a standard thimble ionization chamber (0.125 cm3: PTW, Freiburg, Germany) or a smaller PinPoint chamber (0.015 cm3: PTW), and measurements can be made with either chamber in a range of positions throughout the phantom. Coronal films can also be acquired within the phantom, and additional solid blocks of Perspex allow for transverse films to be acquired within the head region. Initial studies using simple conventional head-and-neck plans established the reproducibility of the phantom and the measurement devices to within the setup uncertainty of +/- 0.5 mm. Subsequent verification of 9 clinical head-and-neck IMRT plans demonstrated the efficacy of the phantom in making a range of patient-specific dose measurements in regions of dosimetric and clinical interest. Agreement between measured values and those predicted by the Pinnacle3 treatment planning system (Philips Medical Systems, Andover, MA) was found to be generally good, with a mean error on the calculated dose to each point of +0.2% (range: -4.3% to +2.2%; n = 9) for the primary planning target volume (PTV), -0.1% (range: -1.5% to +2.0%; n = 8) for the nodal PTV, and +0.0% (range: -1.8% to +4.3%, n = 9) for the spinal cord. The suitability of the phantom for measuring combined dose distributions using radiographic film was also evaluated. The phantom has proved to be a valuable tool in the development and implementation of clinical head-and-neck IMRT, allowing for accurate verification of absolute dose and dose distributions in regions of clinical and dosimetric interest.

  3. Feasibility of radiosurgery for patients with spinal tumors treated in lateral decubitus position: A case series from Memorial Sloan Kettering Cancer Center.

    PubMed

    Navo, Elliot B; Lovelock, D Michael; Zatcky, Joan; Yamada, Josh

    2017-01-01

    Often in clinical practice radiation oncologists encounter patients who require treatment to the spine commonly in the setting of metastatic disease. These metastases usually cause pain, immobility, or neurologic deficits mandating expedited therapy to alleviate the suffering of our patients. Spine radiosurgery techniques have been used extensively for palliation purposes; however, given the patients' deteriorating condition or pain and inability to tolerate anesthesia the radiation oncologist is often left with the conundrum of how to best set up his or her patient in preparation for radiosurgery if supine is not a viable option. In the Memorial Sloan Kettering Cancer Center several patients have been treated successfully in the lateral decubitus position to overcome this set-up issue. In this report, the feasibility of the lateral decubitus set-up for patients who benefit from radiosurgery to the spine when and if they cannot tolerate standard supine position is explored. To report on a retrospective case series of three patients with a total of four lesions who were treated with radiosurgery for spinal metastases while set up in the lateral decubitus position. This is a retrospective case series of 3 patients who were treated with radiosurgery to the spine for palliation of painful metastatic foci. Patients were treated in the lateral decubitus position in 1-5 fractions in order to be eligible for this retrospective case series. Their set-up data, and clinical outcomes were then compared with historic controls. Patients who were treated in the lateral decubitus position were set up reliably and reproducibly. Additionally clinical outcomes on routine follow-up and imaging, and toxicity profiles also corroborated the utility of this treatment set-up. Routinely employing optical surface tracking during patient setup followed by KVCBCT prior to treatment delivery along with intra-fractional monitoring is safe and effective while utilizing the lateral decubitus position for the treatment of spinal metastases for patients who cannot tolerate the supine position. Finally the patient follow-up also corroborated that treatments were successful thus lending credence to the safety, ease, effectiveness, and feasibility of this patient set-up.

  4. Dedicated magnetic resonance imaging in the radiotherapy clinic.

    PubMed

    Karlsson, Mikael; Karlsson, Magnus G; Nyholm, Tufve; Amies, Christopher; Zackrisson, Björn

    2009-06-01

    To introduce a novel technology arrangement in an integrated environment and outline the logistics model needed to incorporate dedicated magnetic resonance (MR) imaging in the radiotherapy workflow. An initial attempt was made to analyze the value and feasibility of MR-only imaging compared to computed tomography (CT) imaging, testing the assumption that MR is a better choice for target and healthy tissue delineation in radiotherapy. A 1.5-T MR unit with a 70-cm-bore size was installed close to a linear accelerator, and a special trolley was developed for transporting patients who were fixated in advance between the MR unit and the accelerator. New MR-based workflow procedures were developed and evaluated. MR-only treatment planning has been facilitated, thus avoiding all registration errors between CT and MR scans, but several new aspects of MR imaging must be considered. Electron density information must be obtained by other methods. Generation of digitally reconstructed radiographs (DRR) for x-ray setup verification is not straight forward, and reliable corrections of geometrical distortions must be applied. The feasibility of MR imaging virtual simulation has been demonstrated, but a key challenge to overcome is correct determination of the skeleton, which is often needed for the traditional approach of beam modeling. The trolley solution allows for a highly precise setup for soft tissue tumors without the invasive handling of radiopaque markers. The new logistics model with an integrated MR unit is efficient and will allow for improved tumor definition and geometrical precision without a significant loss of dosimetric accuracy. The most significant development needed is improved bone imaging.

  5. SU-E-J-90: MRI-Based Treatment Simulation and Patient Setup for Radiation Therapy of Brain Cancer

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Yang, Y; Cao, M; Han, F

    2014-06-01

    Purpose: Traditional radiation therapy of cancer is heavily dependent on CT. CT provides excellent depiction of the bones but lacks good soft tissue contrast, which makes contouring difficult. Often, MRIs are fused with CT to take advantage of its superior soft tissue contrast. Such an approach has drawbacks. It is desirable to perform treatment simulation entirely based on MRI. To achieve MR-based simulation for radiation therapy, bone imaging is an important challenge because of the low MR signal intensity from bone due to its ultra-short T2 and T1, which presents difficulty for both dose calculation and patient setup in termsmore » of digitally reconstructed radiograph (DRR) generation. Current solutions will either require manual bone contouring or multiple MR scans. We present a technique to generate DRR using MRI with an Ultra Short Echo Time (UTE) sequence which is applicable to both OBI and ExacTrac 2D patient setup. Methods: Seven brain cancer patients were scanned at 1.5 Tesla using a radial UTE sequence. The sequence acquires two images at two different echo times. The two images were processed using in-house software. The resultant bone images were subsequently loaded into commercial systems to generate DRRs. Simulation and patient clinical on-board images were used to evaluate 2D patient setup with MRI-DRRs. Results: The majority bones are well visualized in all patients. The fused image of patient CT with the MR bone image demonstrates the accuracy of automatic bone identification using our technique. The generated DRR is of good quality. Accuracy of 2D patient setup by using MRI-DRR is comparable to CT-based 2D patient setup. Conclusion: This study shows the potential of DRR generation with single MR sequence. Further work will be needed on MR sequence development and post-processing procedure to achieve robust MR bone imaging for other human sites in addition to brain.« less

  6. The influence of verification jig on framework fit for nonsegmented fixed implant-supported complete denture.

    PubMed

    Ercoli, Carlo; Geminiani, Alessandro; Feng, Changyong; Lee, Heeje

    2012-05-01

    The purpose of this retrospective study was to assess if there was a difference in the likelihood of achieving passive fit when an implant-supported full-arch prosthesis framework is fabricated with or without the aid of a verification jig. This investigation was approved by the University of Rochester Research Subject Review Board (protocol #RSRB00038482). Thirty edentulous patients, 49 to 73 years old (mean 61 years old), rehabilitated with a nonsegmented fixed implant-supported complete denture were included in the study. During the restorative process, final impressions were made using the pickup impression technique and elastomeric impression materials. For 16 patients, a verification jig was made (group J), while for the remaining 14 patients, a verification jig was not used (group NJ) and the framework was fabricated directly on the master cast. During the framework try-in appointment, the fit was assessed by clinical (Sheffield test) and radiographic inspection and recorded as passive or nonpassive. When a verification jig was used (group J, n = 16), all frameworks exhibited clinically passive fit, while when a verification jig was not used (group NJ, n = 14), only two frameworks fit. This difference was statistically significant (p < .001). Within the limitations of this retrospective study, the fabrication of a verification jig ensured clinically passive fit of metal frameworks in nonsegmented fixed implant-supported complete denture. © 2011 Wiley Periodicals, Inc.

  7. Feasibility of radiosurgery for patients with spinal tumors treated in lateral decubitus position: A case series from Memorial Sloan Kettering Cancer Center

    PubMed Central

    Lovelock, D. Michael; Zatcky, Joan; Yamada, Josh

    2017-01-01

    Introduction Often in clinical practice radiation oncologists encounter patients who require treatment to the spine commonly in the setting of metastatic disease. These metastases usually cause pain, immobility, or neurologic deficits mandating expedited therapy to alleviate the suffering of our patients. Spine radiosurgery techniques have been used extensively for palliation purposes; however, given the patients’ deteriorating condition or pain and inability to tolerate anesthesia the radiation oncologist is often left with the conundrum of how to best set up his or her patient in preparation for radiosurgery if supine is not a viable option. In the Memorial Sloan Kettering Cancer Center several patients have been treated successfully in the lateral decubitus position to overcome this set-up issue. In this report, the feasibility of the lateral decubitus set-up for patients who benefit from radiosurgery to the spine when and if they cannot tolerate standard supine position is explored. Objective To report on a retrospective case series of three patients with a total of four lesions who were treated with radiosurgery for spinal metastases while set up in the lateral decubitus position. Methods and materials This is a retrospective case series of 3 patients who were treated with radiosurgery to the spine for palliation of painful metastatic foci. Patients were treated in the lateral decubitus position in 1-5 fractions in order to be eligible for this retrospective case series. Their set-up data, and clinical outcomes were then compared with historic controls. Results Patients who were treated in the lateral decubitus position were set up reliably and reproducibly. Additionally clinical outcomes on routine follow-up and imaging, and toxicity profiles also corroborated the utility of this treatment set-up. Conclusions Routinely employing optical surface tracking during patient setup followed by KVCBCT prior to treatment delivery along with intra-fractional monitoring is safe and effective while utilizing the lateral decubitus position for the treatment of spinal metastases for patients who cannot tolerate the supine position. Finally the patient follow-up also corroborated that treatments were successful thus lending credence to the safety, ease, effectiveness, and feasibility of this patient set-up. PMID:29296455

  8. Poster - Thurs Eve-43: Verification of dose calculation with tissue inhomogeneity using MapCHECK.

    PubMed

    Korol, R; Chen, J; Mosalaei, H; Karnas, S

    2008-07-01

    MapCHECK (Sun Nuclear, Melbourne, FL) with 445 diode detectors has been used widely for routine IMRT quality assurance (QA) 1 . However, routine IMRT QA has not included the verification of inhomogeneity effects. The objective of this study is to use MapCHECK and a phantom to verify dose calculation and IMRT delivery with tissue inhomogeneity. A phantom with tissue inhomogeneities was placed on top of MapCHECK to measure the planar dose for an anterior beam with photon energy 6 MV or 18 MV. The phantom was composed of a 3.5 cm thick block of lung equivalent material and solid water arranged side by side with a 0.5 cm slab of solid water on the top of the phantom. The phantom setup including MapCHECK was CT scanned and imported into Pinnacle 8.0d for dose calculation. Absolute dose distributions were compared with gamma criteria 3% for dose difference and 3 mm for distance-to-agreement. The results are in good agreement between the measured and calculated planar dose with 88% pass rate based on the gamma analysis. The major dose difference was at the lung-water interface. Further investigation will be performed on a custom designed inhomogeneity phantom with inserts of varying densities and effective depth to create various dose gradients at the interface for dose calculation and delivery verification. In conclusion, a phantom with tissue inhomogeneities can be used with MapCHECK for verification of dose calculation and delivery with tissue inhomogeneity. © 2008 American Association of Physicists in Medicine.

  9. Numerical Weather Predictions Evaluation Using Spatial Verification Methods

    NASA Astrophysics Data System (ADS)

    Tegoulias, I.; Pytharoulis, I.; Kotsopoulos, S.; Kartsios, S.; Bampzelis, D.; Karacostas, T.

    2014-12-01

    During the last years high-resolution numerical weather prediction simulations have been used to examine meteorological events with increased convective activity. Traditional verification methods do not provide the desired level of information to evaluate those high-resolution simulations. To assess those limitations new spatial verification methods have been proposed. In the present study an attempt is made to estimate the ability of the WRF model (WRF -ARW ver3.5.1) to reproduce selected days with high convective activity during the year 2010 using those feature-based verification methods. Three model domains, covering Europe, the Mediterranean Sea and northern Africa (d01), the wider area of Greece (d02) and central Greece - Thessaly region (d03) are used at horizontal grid-spacings of 15km, 5km and 1km respectively. By alternating microphysics (Ferrier, WSM6, Goddard), boundary layer (YSU, MYJ) and cumulus convection (Kain-­-Fritsch, BMJ) schemes, a set of twelve model setups is obtained. The results of those simulations are evaluated against data obtained using a C-Band (5cm) radar located at the centre of the innermost domain. Spatial characteristics are well captured but with a variable time lag between simulation results and radar data. Acknowledgements: This research is co­financed by the European Union (European Regional Development Fund) and Greek national funds, through the action "COOPERATION 2011: Partnerships of Production and Research Institutions in Focused Research and Technology Sectors" (contract number 11SYN_8_1088 - DAPHNE) in the framework of the operational programme "Competitiveness and Entrepreneurship" and Regions in Transition (OPC II, NSRF 2007-­-2013).

  10. Impact of patient-specific factors, irradiated left ventricular volume, and treatment set-up errors on the development of myocardial perfusion defects after radiation therapy for left-sided breast cancer

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Evans, Elizabeth S.; Prosnitz, Robert G.; Yu Xiaoli

    2006-11-15

    Purpose: The aim of this study was to assess the impact of patient-specific factors, left ventricle (LV) volume, and treatment set-up errors on the rate of perfusion defects 6 to 60 months post-radiation therapy (RT) in patients receiving tangential RT for left-sided breast cancer. Methods and Materials: Between 1998 and 2005, a total of 153 patients were enrolled onto an institutional review board-approved prospective study and had pre- and serial post-RT (6-60 months) cardiac perfusion scans to assess for perfusion defects. Of the patients, 108 had normal pre-RT perfusion scans and available follow-up data. The impact of patient-specific factors onmore » the rate of perfusion defects was assessed at various time points using univariate and multivariate analysis. The impact of set-up errors on the rate of perfusion defects was also analyzed using a one-tailed Fisher's Exact test. Results: Consistent with our prior results, the volume of LV in the RT field was the most significant predictor of perfusion defects on both univariate (p = 0.0005 to 0.0058) and multivariate analysis (p = 0.0026 to 0.0029). Body mass index (BMI) was the only significant patient-specific factor on both univariate (p = 0.0005 to 0.022) and multivariate analysis (p = 0.0091 to 0.05). In patients with very small volumes of LV in the planned RT fields, the rate of perfusion defects was significantly higher when the fields set-up 'too deep' (83% vs. 30%, p = 0.059). The frequency of deep set-up errors was significantly higher among patients with BMI {>=}25 kg/m{sup 2} compared with patients of normal weight (47% vs. 28%, p = 0.068). Conclusions: BMI {>=}25 kg/m{sup 2} may be a significant risk factor for cardiac toxicity after RT for left-sided breast cancer, possibly because of more frequent deep set-up errors resulting in the inclusion of additional heart in the RT fields. Further study is necessary to better understand the impact of patient-specific factors and set-up errors on the development of RT-induced perfusion defects.« less

  11. SU-E-T-657: Quantitative Assessment of Plan Robustness for Helical Tomotherapy for Head and Neck Cancer Radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Matney, J; Lian, J; Chera, B

    2015-06-15

    Introduction: Geometric uncertainties in daily patient setup can lead to variations in the planned dose, especially when using highly conformal techniques such as helical Tomotherapy. To account for the potential effect of geometric uncertainty, our clinical practice is to expand critical structures by 3mm expansion into planning risk volumes (PRV). The PRV concept assumes the spatial dose cloud is insensitive to patient positioning. However, no tools currently exist to determine if a Tomotherapy plan is robust to the effects of daily setup variation. We objectively quantified the impact of geometric uncertainties on the 3D doses to critical normal tissues duringmore » helical Tomotherapy. Methods: Using a Matlab-based program created and validated by Accuray (Madison, WI), the planned Tomotherapy delivery sinogram recalculated dose on shifted CT datasets. Ten head and neck patients were selected for analysis. To simulate setup uncertainty, the patient anatomy was shifted ±3mm in the longitudinal, lateral and vertical axes. For each potential shift, the recalculated doses to various critical normal tissues were compared to the doses delivered to the PRV in the original plan Results: 18 shifted scenarios created from Tomotherapy plans for three patients with head and neck cancers were analyzed. For all simulated setup errors, the maximum doses to the brainstem, spinal cord, parotids and cochlea were no greater than 0.6Gy of the respective original PRV maximum. Despite 3mm setup shifts, the minimum dose delivered to 95% of the CTVs and PTVs were always within 0.4Gy of the original plan. Conclusions: For head and neck sites treated with Tomotherapy, the use of a 3mm PRV expansion provide a reasonable estimate of the dosimetric effects of 3mm setup uncertainties. Similarly, target coverage appears minimally effected by a 3mm setup uncertainty. Data from a larger number of patients will be presented. Future work will include other anatomical sites.« less

  12. Tumor control probability reduction in gated radiotherapy of non-small cell lung cancers: a feasibility study.

    PubMed

    Siochi, R Alfredo; Kim, Yusung; Bhatia, Sudershan

    2014-10-16

    We studied the feasibility of evaluating tumor control probability (TCP) reductions for tumor motion beyond planned gated radiotherapy margins. Tumor motion was determined from cone-beam CT projections acquired for patient setup, intrafraction respiratory traces, and 4D CTs for five non-small cell lung cancer (NSCLC) patients treated with gated radiotherapy. Tumors were subdivided into 1 mm sections whose positions and doses were determined for each beam-on time point. (The dose calculation model was verified with motion phantom measurements.) The calculated dose distributions were used to generate the treatment TCPs for each patient. The plan TCPs were calculated from the treatment planning dose distributions. The treatment TCPs were compared to the plan TCPs for various models and parameters. Calculated doses matched phantom measurements within 0.3% for up to 3 cm of motion. TCP reductions for excess motion greater than 5mm ranged from 1.7% to 11.9%, depending on model parameters, and were as high as 48.6% for model parameters that simulated an individual patient. Repeating the worst case motion for all fractions increased TCP reductions by a factor of 2 to 3, while hypofractionation decreased these reductions by as much as a factor of 3. Treatment motion exceeding gating margins by more than 5 mm can lead to considerable TCP reductions. Appropriate margins for excess motion are recommended, unless applying daily tumor motion verification and adjusting thegating window.

  13. Image guided radiation therapy applications for head and neck, prostate, and breast cancers using 3D ultrasound imaging and Monte Carlo dose calculations

    NASA Astrophysics Data System (ADS)

    Fraser, Danielle

    In radiation therapy an uncertainty in the delivered dose always exists because anatomic changes are unpredictable and patient specific. Image guided radiation therapy (IGRT) relies on imaging in the treatment room to monitor the tumour and surrounding tissue to ensure their prescribed position in the radiation beam. The goal of this thesis was to determine the dosimetric impact on the misaligned radiation therapy target for three cancer sites due to common setup errors; organ motion, tumour tissue deformation, changes in body habitus, and treatment planning errors. For this purpose, a novel 3D ultrasound system (Restitu, Resonant Medical, Inc.) was used to acquire a reference image of the target in the computed tomography simulation room at the time of treatment planning, to acquire daily images in the treatment room at the time of treatment delivery, and to compare the daily images to the reference image. The measured differences in position and volume between daily and reference geometries were incorporated into Monte Carlo (MC) dose calculations. The EGSnrc (National Research Council, Canada) family of codes was used to model Varian linear accelerators and patient specific beam parameters, as well as to estimate the dose to the target and organs at risk under several different scenarios. After validating the necessity of MC dose calculations in the pelvic region, the impact of interfraction prostate motion, and subsequent patient realignment under the treatment beams, on the delivered dose was investigated. For 32 patients it is demonstrated that using 3D conformal radiation therapy techniques and a 7 mm margin, the prescribed dose to the prostate, rectum, and bladder is recovered within 0.5% of that planned when patient setup is corrected for prostate motion, despite the beams interacting with a new external surface and internal tissue boundaries. In collaboration with the manufacturer, the ultrasound system was adapted from transabdominal imaging to neck imaging. Two case studies of nasopharyngeal cancer are discussed. The deformation of disease-positive cervical lymph nodes was monitored throughout treatment. Node volumes shrunk to 17% of the initial volume, moved up 1.3 cm, and received up to a 12% lower dose than that prescribed. It is shown that difficulties in imaging soft tissue in the neck region are circumvented with ultrasound imaging, and after dosimetric verification it is argued that adaptive replanning may be more beneficial than patient realignment when intensity modulated radiation therapy techniques are used. Some of the largest dose delivery errors were found in external electron beam treatments for breast cancer patients who underwent breast conserving surgery. Inaccuracies in conventional treatment planning resulted in substantial target dose discrepancies of up to 88%. When patient setup errors, interfraction tumour bed motion, and tissue remodeling were considered, inadequate target coverage was exacerbated. This thesis quantifies the dose discrepancy between that prescribed and that delivered. I delve into detail for common IGRT treatment sites, and illuminate problems that have not received much attention for less common IGRT treatment sites.

  14. Response to "Improving Patient Safety With Error Identification in Chemotherapy Orders by Verification Nurses"
.

    PubMed

    Zhu, Ling-Ling; Lv, Na; Zhou, Quan

    2016-12-01

    We read, with great interest, the study by Baldwin and Rodriguez (2016), which described the role of the verification nurse and details the verification process in identifying errors related to chemotherapy orders. We strongly agree with their findings that a verification nurse, collaborating closely with the prescribing physician, pharmacist, and treating nurse, can better identify errors and maintain safety during chemotherapy administration.

  15. Site‐specific tolerance tables and indexing device to improve patient setup reproducibility

    PubMed Central

    James, Joshua A.; Cetnar, Ashley J.; McCullough, Mark A.; Wang, Brian

    2015-01-01

    While the implementation of tools such as image‐guidance and immobilization devices have helped to prevent geometric misses in radiation therapy, many treatments remain prone to error if these items are not available, not utilized for every fraction, or are misused. The purpose of this project is to design a set of site‐specific treatment tolerance tables to be applied to the treatment couch for use in a record and verify (R&V) system that will insure accurate patient setup with minimal workflow interruption. This project also called for the construction of a simple indexing device to help insure reproducible patient setup for patients that could not be indexed with existing equipment. The tolerance tables were created by retrospective analysis on a total of 66 patients and 1,308 treatments, separating them into five categories based on disease site: lung, head and neck (H&N), breast, pelvis, and abdomen. Couch parameter tolerance tables were designed to encompass 95% of treatments, and were generated by calculating the standard deviation of couch vertical, longitudinal, and lateral values using the first day of treatment as a baseline. We also investigated an alternative method for generating the couch tolerances by updating the baseline values when patient position was verified with image guidance. This was done in order to adapt the tolerances to any gradual changes in patient setup that would not correspond with a mistreatment. The tolerance tables and customizable indexing device were then implemented for a trial period in order to determine the feasibility of the system. During this trial period we collected data from 1,054 fractions from 65 patients. We then analyzed the number of treatments that would have been out of tolerance, as well as whether or not the tolerances or setup techniques should be adjusted. When the couch baseline values were updated with every imaging fraction, the average rate of tolerance violations was 10% for the lung, H&N, abdomen, and pelvis treatments. Using the indexing device, tolerances for patients with pelvic disease decreased (e.g., from 5.3 cm to 4.3 cm longitudinally). Unfortunately, the results from breast patients were highly variable due to the complexity of the setup technique, making the couch an inadequate surrogate for measuring setup accuracy. In summary, we have developed a method to turn the treatment couch parameters within the R&V system into a useful alert tool, which can be implemented at other institutions, in order to identify potential errors in patient setup. PACS numbers: 87.53Kn, 87.55.kh, 87.55.ne, 87.55.km, 87.55K‐, 87.55.Qr PMID:26103475

  16. First clinical experience in carbon ion scanning beam therapy: retrospective analysis of patient positional accuracy.

    PubMed

    Mori, Shinichiro; Shibayama, Kouichi; Tanimoto, Katsuyuki; Kumagai, Motoki; Matsuzaki, Yuka; Furukawa, Takuji; Inaniwa, Taku; Shirai, Toshiyuki; Noda, Koji; Tsuji, Hiroshi; Kamada, Tadashi

    2012-09-01

    Our institute has constructed a new treatment facility for carbon ion scanning beam therapy. The first clinical trials were successfully completed at the end of November 2011. To evaluate patient setup accuracy, positional errors between the reference Computed Tomography (CT) scan and final patient setup images were calculated using 2D-3D registration software. Eleven patients with tumors of the head and neck, prostate and pelvis receiving carbon ion scanning beam treatment participated. The patient setup process takes orthogonal X-ray flat panel detector (FPD) images and the therapists adjust the patient table position in six degrees of freedom to register the reference position by manual or auto- (or both) registration functions. We calculated residual positional errors with the 2D-3D auto-registration function using the final patient setup orthogonal FPD images and treatment planning CT data. Residual error averaged over all patients in each fraction decreased from the initial to the last treatment fraction [1.09 mm/0.76° (averaged in the 1st and 2nd fractions) to 0.77 mm/0.61° (averaged in the 15th and 16th fractions)]. 2D-3D registration calculation time was 8.0 s on average throughout the treatment course. Residual errors in translation and rotation averaged over all patients as a function of date decreased with the passage of time (1.6 mm/1.2° in May 2011 to 0.4 mm/0.2° in December 2011). This retrospective residual positional error analysis shows that the accuracy of patient setup during the first clinical trials of carbon ion beam scanning therapy was good and improved with increasing therapist experience.

  17. A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy.

    PubMed

    Boswell, Sarah A; Jeraj, Robert; Ruchala, Kenneth J; Olivera, Gustavo H; Jaradat, Hazim A; James, Joshua A; Gutierrez, Alonso; Pearson, Dave; Frank, Gary; Mackie, T Rock

    2005-06-01

    An accurate means of determining and correcting for daily patient setup errors is important to the cancer outcome in radiotherapy. While many tools have been developed to detect setup errors, difficulty may arise in accurately adjusting the patient to account for the rotational error components. A novel, automated method to correct for rotational patient setup errors in helical tomotherapy is proposed for a treatment couch that is restricted to motion along translational axes. In tomotherapy, only a narrow superior/inferior section of the target receives a dose at any instant, thus rotations in the sagittal and coronal planes may be approximately corrected for by very slow continuous couch motion in a direction perpendicular to the scanning direction. Results from proof-of-principle tests indicate that the method improves the accuracy of treatment delivery, especially for long and narrow targets. Rotational corrections about an axis perpendicular to the transverse plane continue to be implemented easily in tomotherapy by adjustment of the initial gantry angle.

  18. SU-D-BRC-03: Development and Validation of an Online 2D Dose Verification System for Daily Patient Plan Delivery Accuracy Check

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zhao, J; Hu, W; Xing, Y

    Purpose: All plan verification systems for particle therapy are designed to do plan verification before treatment. However, the actual dose distributions during patient treatment are not known. This study develops an online 2D dose verification tool to check the daily dose delivery accuracy. Methods: A Siemens particle treatment system with a modulated scanning spot beam is used in our center. In order to do online dose verification, we made a program to reconstruct the delivered 2D dose distributions based on the daily treatment log files and depth dose distributions. In the log files we can get the focus size, positionmore » and particle number for each spot. A gamma analysis is used to compare the reconstructed dose distributions with the dose distributions from the TPS to assess the daily dose delivery accuracy. To verify the dose reconstruction algorithm, we compared the reconstructed dose distributions to dose distributions measured using PTW 729XDR ion chamber matrix for 13 real patient plans. Then we analyzed 100 treatment beams (58 carbon and 42 proton) for prostate, lung, ACC, NPC and chordoma patients. Results: For algorithm verification, the gamma passing rate was 97.95% for the 3%/3mm and 92.36% for the 2%/2mm criteria. For patient treatment analysis,the results were 97.7%±1.1% and 91.7%±2.5% for carbon and 89.9%±4.8% and 79.7%±7.7% for proton using 3%/3mm and 2%/2mm criteria, respectively. The reason for the lower passing rate for the proton beam is that the focus size deviations were larger than for the carbon beam. The average focus size deviations were −14.27% and −6.73% for proton and −5.26% and −0.93% for carbon in the x and y direction respectively. Conclusion: The verification software meets our requirements to check for daily dose delivery discrepancies. Such tools can enhance the current treatment plan and delivery verification processes and improve safety of clinical treatments.« less

  19. Current status of 3D EPID-based in vivo dosimetry in The Netherlands Cancer Institute

    NASA Astrophysics Data System (ADS)

    Mijnheer, B.; Olaciregui-Ruiz, I.; Rozendaal, R.; Spreeuw, H.; van Herk, M.; Mans, A.

    2015-01-01

    3D in vivo dose verification using a-Si EPIDs is performed routinely in our institution for almost all RT treatments. The EPID-based 3D dose distribution is reconstructed using a back-projection algorithm and compared with the planned dose distribution using 3D gamma evaluation. Dose-reconstruction and gamma-evaluation software runs automatically, and deviations outside the alert criteria are immediately available and investigated, in combination with inspection of cone-beam CT scans. The implementation of our 3D EPID- based in vivo dosimetry approach was able to replace pre-treatment verification for more than 90% of the patient treatments. Clinically relevant deviations could be detected for approximately 1 out of 300 patient treatments (IMRT and VMAT). Most of these errors were patient related anatomical changes or deviations from the routine clinical procedure, and would not have been detected by pre-treatment verification. Moreover, 3D EPID-based in vivo dose verification is a fast and accurate tool to assure the safe delivery of RT treatments. It provides clinically more useful information and is less time consuming than pre-treatment verification measurements. Automated 3D in vivo dosimetry is therefore a prerequisite for large-scale implementation of patient-specific quality assurance of RT treatments.

  20. SU-E-J-39: Dosimetric Benefit of Implanted Marker-Based CBCT Setup for Definitive Prostatic Radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zhen, H; Wu, Z; Bluemenfeld, P

    Purpose Daily setup for definitive prostatic radiotherapy is challenged by suboptimal visibility of the prostate boundary and daily variation of rectum shape and position. For patients with improved bowel preparation, we conducted a dosimetric comparison between prostate implanted marker (IM)-based daily setup and anterior rectal wall (ARW)-based setup, with the hypothesis that the former leads to adequate target coverage with better rectal sparing. Methods Five IMRT/VMAT prostate cases with implanted markers were selected for analysis. Daily CBCT showed improvement of the rectal volume compared to planning CT. For each patient, the prostate and rectum were contoured on three CBCT imagesmore » (fraction 5/15/25) with subsequent physician review. The CBCTs were then registered to a planning CT using IM-based registration. The deviation of ARW positions from planning CT to CBCT were analyzed at various sup-inf levels (−1.8 cm to 1.8 cm from level of prostate center). To estimate the potential dosimetric impact from ARW-based setup, the treatment plans were recalculated using A-P shifts ranging from −1mm to +6mm. Clinically important rectum DVH values including Dmax, D3cc and Dmean were computed. Results For the studied patients, we observed on average 32% rectum volume reduction from planning CT to CBCT. As a Results, the ARW on average shifts posteriorly by −1mm to +5mm, depending on the sup-inf level of observation, with larger shifts observed at more superior levels. Recalculation shows that when ARW shifts 1mm posteriorly, ARW-based CBCT setup leads to a 1.0%, 4.2%, and 3.2% increase in rectum Dmax, D3cc, and Dmean, respectively, compared to IM-based setup. The dosimetric deviations increase to 4.7%, 25.8% and 24.7% when ARW shifts 6mm posteriorly. No significant prostate-only dose difference was observed. Conclusion For patients with improved bowel preparation, IM-based CBCT setup leads to accurate prostate coverage along with significantly lower rectal dose, compared to ARW-based setup.« less

  1. Adjusting for partial verification or workup bias in meta-analyses of diagnostic accuracy studies.

    PubMed

    de Groot, Joris A H; Dendukuri, Nandini; Janssen, Kristel J M; Reitsma, Johannes B; Brophy, James; Joseph, Lawrence; Bossuyt, Patrick M M; Moons, Karel G M

    2012-04-15

    A key requirement in the design of diagnostic accuracy studies is that all study participants receive both the test under evaluation and the reference standard test. For a variety of practical and ethical reasons, sometimes only a proportion of patients receive the reference standard, which can bias the accuracy estimates. Numerous methods have been described for correcting this partial verification bias or workup bias in individual studies. In this article, the authors describe a Bayesian method for obtaining adjusted results from a diagnostic meta-analysis when partial verification or workup bias is present in a subset of the primary studies. The method corrects for verification bias without having to exclude primary studies with verification bias, thus preserving the main advantages of a meta-analysis: increased precision and better generalizability. The results of this method are compared with the existing methods for dealing with verification bias in diagnostic meta-analyses. For illustration, the authors use empirical data from a systematic review of studies of the accuracy of the immunohistochemistry test for diagnosis of human epidermal growth factor receptor 2 status in breast cancer patients.

  2. Assessment of image quality of a radiotherapy-specific hardware solution for PET/MRI in head and neck cancer patients.

    PubMed

    Winter, René M; Leibfarth, Sara; Schmidt, Holger; Zwirner, Kerstin; Mönnich, David; Welz, Stefan; Schwenzer, Nina F; la Fougère, Christian; Nikolaou, Konstantin; Gatidis, Sergios; Zips, Daniel; Thorwarth, Daniela

    2018-05-07

    Functional PET/MRI has great potential to improve radiotherapy planning (RTP). However, data integration requires imaging with radiotherapy-specific patient positioning. Here, we investigated the feasibility and image quality of radiotherapy-customized PET/MRI in head-and-neck cancer (HNC) patients using a dedicated hardware setup. Ten HNC patients were examined with simultaneous PET/MRI before treatment, with radiotherapy and diagnostic scan setup, respectively. We tested feasibility of radiotherapy-specific patient positioning and compared the image quality between both setups by pairwise image analysis of 18 F-FDG-PET, T1/T2-weighted and diffusion-weighted MRI. For image quality assessment, similarity measures including average symmetric surface distance (ASSD) of PET and MR-based tumor contours, MR signal-to-noise ratio (SNR) and mean apparent diffusion coefficient (ADC) value were used. PET/MRI in radiotherapy position was feasible - all patients were successfully examined. ASSD (median/range) of PET and MR contours was 0.6 (0.4-1.2) and 0.9 (0.5-1.3) mm, respectively. For T2-weighted MRI, a reduced SNR of -26.2% (-39.0--11.7) was observed with radiotherapy setup. No significant difference in mean ADC was found. Simultaneous PET/MRI in HNC patients using radiotherapy positioning aids is clinically feasible. Though SNR was reduced, the image quality obtained with a radiotherapy setup meets RTP requirements and the data can thus be used for personalized RTP. Copyright © 2018 The Author(s). Published by Elsevier B.V. All rights reserved.

  3. Patient-specific IMRT verification using independent fluence-based dose calculation software: experimental benchmarking and initial clinical experience.

    PubMed

    Georg, Dietmar; Stock, Markus; Kroupa, Bernhard; Olofsson, Jörgen; Nyholm, Tufve; Ahnesjö, Anders; Karlsson, Mikael

    2007-08-21

    Experimental methods are commonly used for patient-specific intensity-modulated radiotherapy (IMRT) verification. The purpose of this study was to investigate the accuracy and performance of independent dose calculation software (denoted as 'MUV' (monitor unit verification)) for patient-specific quality assurance (QA). 52 patients receiving step-and-shoot IMRT were considered. IMRT plans were recalculated by the treatment planning systems (TPS) in a dedicated QA phantom, in which an experimental 1D and 2D verification (0.3 cm(3) ionization chamber; films) was performed. Additionally, an independent dose calculation was performed. The fluence-based algorithm of MUV accounts for collimator transmission, rounded leaf ends, tongue-and-groove effect, backscatter to the monitor chamber and scatter from the flattening filter. The dose calculation utilizes a pencil beam model based on a beam quality index. DICOM RT files from patient plans, exported from the TPS, were directly used as patient-specific input data in MUV. For composite IMRT plans, average deviations in the high dose region between ionization chamber measurements and point dose calculations performed with the TPS and MUV were 1.6 +/- 1.2% and 0.5 +/- 1.1% (1 S.D.). The dose deviations between MUV and TPS slightly depended on the distance from the isocentre position. For individual intensity-modulated beams (total 367), an average deviation of 1.1 +/- 2.9% was determined between calculations performed with the TPS and with MUV, with maximum deviations up to 14%. However, absolute dose deviations were mostly less than 3 cGy. Based on the current results, we aim to apply a confidence limit of 3% (with respect to the prescribed dose) or 6 cGy for routine IMRT verification. For off-axis points at distances larger than 5 cm and for low dose regions, we consider 5% dose deviation or 10 cGy acceptable. The time needed for an independent calculation compares very favourably with the net time for an experimental approach. The physical effects modelled in the dose calculation software MUV allow accurate dose calculations in individual verification points. Independent calculations may be used to replace experimental dose verification once the IMRT programme is mature.

  4. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Latty, Drew, E-mail: drew.latty@health.nsw.gov.au; Stuart, Kirsty E; Westmead Breast Cancer Institute, Sydney, New South Wales

    Radiation treatment to the left breast is associated with increased cardiac morbidity and mortality. The deep inspiration breath-hold technique (DIBH) can decrease radiation dose delivered to the heart and this may facilitate the treatment of the internal mammary chain nodes. The aim of this review is to critically analyse the literature available in relation to breath-hold methods, implementation, utilisation, patient compliance, planning methods and treatment verification of the DIBH technique. Despite variation in the literature regarding the DIBH delivery method, patient coaching, visual feedback mechanisms and treatment verification, all methods of DIBH delivery reduce radiation dose to the heart. Furthermore » research is required to determine optimum protocols for patient training and treatment verification to ensure the technique is delivered successfully.« less

  5. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Moura, Eduardo S., E-mail: emoura@wisc.edu; Micka, John A.; Hammer, Cliff G.

    Purpose: This work presents the development of a phantom to verify the treatment planning system (TPS) algorithms used for high-dose-rate (HDR) brachytherapy. It is designed to measure the relative dose in a heterogeneous media. The experimental details used, simulation methods, and comparisons with a commercial TPS are also provided. Methods: To simulate heterogeneous conditions, four materials were used: Virtual Water™ (VM), BR50/50™, cork, and aluminum. The materials were arranged in 11 heterogeneity configurations. Three dosimeters were used to measure the relative response from a HDR {sup 192}Ir source: TLD-100™, Gafchromic{sup ®} EBT3 film, and an Exradin™ A1SL ionization chamber. Tomore » compare the results from the experimental measurements, the various configurations were modeled in the PENELOPE/penEasy Monte Carlo code. Images of each setup geometry were acquired from a CT scanner and imported into BrachyVision™ TPS software, which includes a grid-based Boltzmann solver Acuros™. The results of the measurements performed in the heterogeneous setups were normalized to the dose values measured in the homogeneous Virtual Water™ setup and the respective differences due to the heterogeneities were considered. Additionally, dose values calculated based on the American Association of Physicists in Medicine-Task Group 43 formalism were compared to dose values calculated with the Acuros™ algorithm in the phantom. Calculated doses were compared at the same points, where measurements have been performed. Results: Differences in the relative response as high as 11.5% were found from the homogeneous setup when the heterogeneous materials were inserted into the experimental phantom. The aluminum and cork materials produced larger differences than the plastic materials, with the BR50/50™ material producing results similar to the Virtual Water™ results. Our experimental methods agree with the PENELOPE/penEasy simulations for most setups and dosimeters. The TPS relative differences with the Acuros™ algorithm were similar in both experimental and simulated setups. The discrepancy between the BrachyVision™, Acuros™, and TG-43 dose responses in the phantom described by this work exceeded 12% for certain setups. Conclusions: The results derived from the phantom measurements show good agreement with the simulations and TPS calculations, using Acuros™ algorithm. Differences in the dose responses were evident in the experimental results when heterogeneous materials were introduced. These measurements prove the usefulness of the heterogeneous phantom for verification of HDR treatment planning systems based on model-based dose calculation algorithms.« less

  6. Verifying Data Integrity of Electronically Scanned Pressure Systems at the NASA Glenn Research Center

    NASA Technical Reports Server (NTRS)

    Panek, Joseph W.

    2001-01-01

    The proper operation of the Electronically Scanned Pressure (ESP) System critical to accomplish the following goals: acquisition of highly accurate pressure data for the development of aerospace and commercial aviation systems and continuous confirmation of data quality to avoid costly, unplanned, repeat wind tunnel or turbine testing. Standard automated setup and checkout routines are necessary to accomplish these goals. Data verification and integrity checks occur at three distinct stages, pretest pressure tubing and system checkouts, daily system validation and in-test confirmation of critical system parameters. This paper will give an overview of the existing hardware, software and methods used to validate data integrity.

  7. Mechanical verification of soft-tissue attachment on bioactive glasses and titanium implants.

    PubMed

    Zhao, Desheng; Moritz, Niko; Vedel, Erik; Hupa, Leena; Aro, Hannu T

    2008-07-01

    Soft-tissue attachment is a desired feature of many clinical biomaterials. The aim of the current study was to design a suitable experimental method for tensile testing of implant incorporation with soft-tissues. Conical implants were made of three compositions of bioactive glass (SiO(2)-P(2)O(5)-B(2)O(3)-Na(2)O-K(2)O-CaO-MgO) or titanium fiber mesh (porosity 84.7%). The implants were surgically inserted into the dorsal subcutaneous soft-tissue or back muscles in the rat. Soft-tissue attachment was evaluated by pull-out testing using a custom-made jig 8 weeks after implantation. Titanium fiber mesh implants had developed a relatively high pull-out force in subcutaneous tissue (12.33+/-5.29 N, mean+/-SD) and also measurable attachment with muscle tissue (2.46+/-1.33 N). The bioactive glass implants failed to show mechanically relevant soft-tissue bonding. The experimental set-up of mechanical testing seems to be feasible for verification studies of soft-tissue attachment. The inexpensive small animal model is beneficial for large-scale in vivo screening of new biomaterials.

  8. SU-E-T-261: Development of An Automated System to Detect Patient Identification and Positioning Errors Prior to Radiotherapy Treatment

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jani, S; Low, D; Lamb, J

    2015-06-15

    Purpose: To develop a system that can automatically detect patient identification and positioning errors using 3D computed tomography (CT) setup images and kilovoltage CT (kVCT) planning images. Methods: Planning kVCT images were collected for head-and-neck (H&N), pelvis, and spine treatments with corresponding 3D cone-beam CT (CBCT) and megavoltage CT (MVCT) setup images from TrueBeam and TomoTherapy units, respectively. Patient identification errors were simulated by registering setup and planning images from different patients. Positioning errors were simulated by misaligning the setup image by 1cm to 5cm in the six anatomical directions for H&N and pelvis patients. Misalignments for spine treatments weremore » simulated by registering the setup image to adjacent vertebral bodies on the planning kVCT. A body contour of the setup image was used as an initial mask for image comparison. Images were pre-processed by image filtering and air voxel thresholding, and image pairs were assessed using commonly-used image similarity metrics as well as custom -designed metrics. A linear discriminant analysis classifier was trained and tested on the datasets, and misclassification error (MCE), sensitivity, and specificity estimates were generated using 10-fold cross validation. Results: Our workflow produced MCE estimates of 0.7%, 1.7%, and 0% for H&N, pelvis, and spine TomoTherapy images, respectively. Sensitivities and specificities ranged from 98.0% to 100%. MCEs of 3.5%, 2.3%, and 2.1% were obtained for TrueBeam images of the above sites, respectively, with sensitivity and specificity estimates between 96.2% and 98.4%. MCEs for 1cm H&N/pelvis misalignments were 1.3/5.1% and 9.1/8.6% for TomoTherapy and TrueBeam images, respectively. 2cm MCE estimates were 0.4%/1.6% and 3.1/3.2%, respectively. Vertebral misalignment MCEs were 4.8% and 4.9% for TomoTherapy and TrueBeam images, respectively. Conclusion: Patient identification and gross misalignment errors can be robustly and automatically detected using 3D setup images of two imaging modalities across three commonly-treated anatomical sites.« less

  9. Dosimetric verification for primary focal hypermetabolism of nasopharyngeal carcinoma patients treated with dynamic intensity-modulated radiation therapy.

    PubMed

    Xin, Yong; Wang, Jia-Yang; Li, Liang; Tang, Tian-You; Liu, Gui-Hong; Wang, Jian-She; Xu, Yu-Mei; Chen, Yong; Zhang, Long-Zhen

    2012-01-01

    To make sure the feasibility with (18F)FDG PET/CT to guided dynamic intensity-modulated radiation therapy (IMRT) for nasopharyngeal carcinoma patients, by dosimetric verification before treatment. Chose 11 patients in III~IVA nasopharyngeal carcinoma treated with functional image-guided IMRT and absolute and relative dosimetric verification by Varian 23EX LA, ionization chamber, 2DICA of I'mRT Matrixx and IBA detachable phantom. Drawing outline and making treatment plan were by different imaging techniques (CT and (18F)FDG PET/CT). The dose distributions of the various regional were realized by SMART. The absolute mean errors of interest area were 2.39%±0.66 using 0.6 cc ice chamber. Results using DTA method, the average relative dose measurements within our protocol (3%, 3 mm) were 87.64% at 300 MU/min in all filed. Dosimetric verification before IMRT is obligatory and necessary. Ionization chamber and 2DICA of I'mRT Matrixx was the effective dosimetric verification tool for primary focal hyper metabolism in functional image-guided dynamic IMRT for nasopharyngeal carcinoma. Our preliminary evidence indicates that functional image-guided dynamic IMRT is feasible.

  10. CFD study on the effects of boundary conditions on air flow through an air-cooled condenser

    NASA Astrophysics Data System (ADS)

    Sumara, Zdeněk; Šochman, Michal

    2018-06-01

    This study focuses on the effects of boundary conditions on effectiveness of an air-cooled condenser (ACC). Heat duty of ACC is very often calculated for ideal uniform velocity field which does not correspond to reality. Therefore, this study studies the effect of wind and different landscapes on air flow through ACC. For this study software OpenFOAM was used and the flow was simulated with the use of RANS equations. For verification of numerical setup a model of one ACC cell with dimensions of platform 1.5×1.5 [m] was used. In this experiment static pressures behind fan and air flows through a model of surface of condenser for different rpm of fan were measured. In OpenFOAM software a virtual clone of this experiment was built and different meshes, turbulent models and numerical schemes were tested. After tuning up numerical setup virtual model of real ACC system was built. Influence of wind, landscape and height of ACC on air flow through ACC has been investigated.

  11. Performing a Large-Scale Modal Test on the B2 Stand Crane at NASA's Stennis Space Center

    NASA Technical Reports Server (NTRS)

    Stasiunas, Eric C.; Parks, Russel A.

    2018-01-01

    A modal test of NASA’s Space Launch System (SLS) Core Stage is scheduled to occur prior to propulsion system verification testing at the Stennis Space Center B2 test stand. A derrick crane with a 180-ft long boom, located at the top of the stand, will be used to suspend the Core Stage in order to achieve defined boundary conditions. During this suspended modal test, it is expected that dynamic coupling will occur between the crane and the Core Stage. Therefore, a separate modal test was performed on the B2 crane itself, in order to evaluate the varying dynamic characteristics and correlate math models of the crane. Performing a modal test on such a massive structure was challenging and required creative test setup and procedures, including implementing both AC and DC accelerometers, and performing both classical hammer and operational modal analysis. This paper describes the logistics required to perform this large-scale test, as well as details of the test setup, the modal test methods used, and an overview of the results.

  12. Verification of endotracheal intubation in obese patients - temporal comparison of ultrasound vs. auscultation and capnography.

    PubMed

    Pfeiffer, P; Bache, S; Isbye, D L; Rudolph, S S; Rovsing, L; Børglum, J

    2012-05-01

    Ultrasound (US) may have an emerging role as an adjunct in verification of endotracheal intubation. Obtaining optimal US images in obese patients is generally regarded more difficult than for other patients. This study compared the time consumption of bilateral lung US with auscultation and capnography for verifying endotracheal intubation in obese patients. A prospective, paired and investigator-blinded study performed in the operating theatre. Twenty-four adult patients requiring endotracheal intubation for bariatric surgery were included. During post-intubation bag ventilation, bilateral lung US was performed for detection of lungsliding indicating lung ventilation simultaneous with capnography and auscultation of epigastrium and chest. Primary outcome measure was the time difference to confirmed endotracheal intubation between US and auscultation alone. The secondary outcome measure was time difference between US and auscultation combined with capnography. Both methods verified endotracheal tube placement in all patients. No significant difference was found between US compared with auscultation alone. Median time for verification by auscultation alone was 47.5 s [interquartile (IQR) 40-51 s], with a mean difference of -0.3 s in favor of US (95% confidence interval -3.5-2.9 s) P = 0.87. Comparing US with the combination of auscultation and capnography, there was a significant difference between the two methods. Median time for verification by US was 43 s (IQR 40-51 s) vs. 55 s (IQR 46-65 s), P < 0.0001. In obese patients, verification of endotracheal tube placement with US is as fast as auscultation alone and faster than the standard method of auscultation and capnography. © 2012 The Authors. Acta Anaesthesiologica Scandinavica © 2012 The Acta Anaesthesiologica Scandinavica Foundation.

  13. Evaluation of patient safety culture among Malaysian retail pharmacists: results of a self-reported survey

    PubMed Central

    Sivanandy, Palanisamy; Maharajan, Mari Kannan; Rajiah, Kingston; Wei, Tan Tyng; Loon, Tan Wee; Yee, Lim Chong

    2016-01-01

    Background Patient safety is a major public health issue, and the knowledge, skills, and experience of health professionals are very much essential for improving patient safety. Patient safety and medication error are very much associated. Pharmacists play a significant role in patient safety. The function of pharmacists in the medication use process is very different from medical and nursing colleagues. Medication dispensing accuracy is a vital element to ensure the safety and quality of medication use. Objective To evaluate the attitude and perception of the pharmacist toward patient safety in retail pharmacies setup in Malaysia. Methods A Pharmacy Survey on Patient Safety Culture questionnaire was used to assess patient safety culture, developed by the Agency for Healthcare Research and Quality, and the convenience sampling method was adopted. Results The overall positive response rate ranged from 31.20% to 87.43%, and the average positive response rate was found to be 67%. Among all the eleven domains pertaining to patient safety culture, the scores of “staff training and skills” were less. Communication openness, and patient counseling are common, but not practiced regularly in the Malaysian retail pharmacy setup compared with those in USA. The overall perception of patient safety of an acceptable level in the current retail pharmacy setup. Conclusion The study revealed that staff training, skills, communication in patient counseling, and communication across shifts and about mistakes are less in current retail pharmacy setup. The overall perception of patient safety should be improved by educating the pharmacists about the significance and essential of patient safety. PMID:27524887

  14. Evaluation of patient safety culture among Malaysian retail pharmacists: results of a self-reported survey.

    PubMed

    Sivanandy, Palanisamy; Maharajan, Mari Kannan; Rajiah, Kingston; Wei, Tan Tyng; Loon, Tan Wee; Yee, Lim Chong

    2016-01-01

    Patient safety is a major public health issue, and the knowledge, skills, and experience of health professionals are very much essential for improving patient safety. Patient safety and medication error are very much associated. Pharmacists play a significant role in patient safety. The function of pharmacists in the medication use process is very different from medical and nursing colleagues. Medication dispensing accuracy is a vital element to ensure the safety and quality of medication use. To evaluate the attitude and perception of the pharmacist toward patient safety in retail pharmacies setup in Malaysia. A Pharmacy Survey on Patient Safety Culture questionnaire was used to assess patient safety culture, developed by the Agency for Healthcare Research and Quality, and the convenience sampling method was adopted. The overall positive response rate ranged from 31.20% to 87.43%, and the average positive response rate was found to be 67%. Among all the eleven domains pertaining to patient safety culture, the scores of "staff training and skills" were less. Communication openness, and patient counseling are common, but not practiced regularly in the Malaysian retail pharmacy setup compared with those in USA. The overall perception of patient safety of an acceptable level in the current retail pharmacy setup. The study revealed that staff training, skills, communication in patient counseling, and communication across shifts and about mistakes are less in current retail pharmacy setup. The overall perception of patient safety should be improved by educating the pharmacists about the significance and essential of patient safety.

  15. Optimizing Cone Beam Computed Tomography (CBCT) System for Image Guided Radiation Therapy

    NASA Astrophysics Data System (ADS)

    Park, Chun Joo

    Cone Beam Computed Tomography (CBCT) system is the most widely used imaging device in image guided radiation therapy (IGRT), where set of 3D volumetric image of patient can be reconstructed to identify and correct position setup errors prior to the radiation treatment. This CBCT system can significantly improve precision of on-line setup errors of patient position and tumor target localization prior to the treatment. However, there are still a number of issues that needs to be investigated with CBCT system such as 1) progressively increasing defective pixels in imaging detectors by its frequent usage, 2) hazardous radiation exposure to patients during the CBCT imaging, 3) degradation of image quality due to patients' respiratory motion when CBCT is acquired and 4) unknown knowledge of certain anatomical features such as liver, due to lack of soft-tissue contrast which makes tumor motion verification challenging. In this dissertation, we explore on optimizing the use of cone beam computed tomography (CBCT) system under such circumstances. We begin by introducing general concept of IGRT. We then present the development of automated defective pixel detection algorithm for X-ray imagers that is used for CBCT imaging using wavelet analysis. We next investigate on developing fast and efficient low-dose volumetric reconstruction techniques which includes 1) fast digital tomosynthesis reconstruction using general-purpose graphics processing unit (GPGPU) programming and 2) fast low-dose CBCT image reconstruction based on the Gradient-Projection-Barzilai-Borwein formulation (GP-BB). We further developed two efficient approaches that could reduce the degradation of CBCT images from respiratory motion. First, we propose reconstructing four dimensional (4D) CBCT and DTS using respiratory signal extracted from fiducial markers implanted in liver. Second, novel motion-map constrained image reconstruction (MCIR) is proposed that allows reconstruction of high quality and high phase resolution 4DCBCT images with no more than the imaging dose used in a standard Free Breathing 3DCBCT (FB-3DCBCT) scan. Finally, we demonstrate a method to analyze motion characteristics of liver that are particularly important for image guided stereotactic body radiation therapy (IG-SBRT). It is anticipated that all the approaches proposed in this study, which are both technically and clinically feasible, will allow much improvement in IGRT process.

  16. SU-E-T-398: Feasibility of Automated Tools for Robustness Evaluation of Advanced Photon and Proton Techniques in Oropharyngeal Cancer

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Liu, H; Liang, X; Kalbasi, A

    2014-06-01

    Purpose: Advanced radiotherapy (RT) techniques such as proton pencil beam scanning (PBS) and photon-based volumetric modulated arc therapy (VMAT) have dosimetric advantages in the treatment of head and neck malignancies. However, anatomic or alignment changes during treatment may limit robustness of PBS and VMAT plans. We assess the feasibility of automated deformable registration tools for robustness evaluation in adaptive PBS and VMAT RT of oropharyngeal cancer (OPC). Methods: We treated 10 patients with bilateral OPC with advanced RT techniques and obtained verification CT scans with physician-reviewed target and OAR contours. We generated 3 advanced RT plans for each patient: protonmore » PBS plan using 2 posterior oblique fields (2F), proton PBS plan using an additional third low-anterior field (3F), and a photon VMAT plan using 2 arcs (Arc). For each of the planning techniques, we forward calculated initial (Ini) plans on the verification scans to create verification (V) plans. We extracted DVH indicators based on physician-generated contours for 2 target and 14 OAR structures to investigate the feasibility of two automated tools (contour propagation (CP) and dose deformation (DD)) as surrogates for routine clinical plan robustness evaluation. For each verification scan, we compared DVH indicators of V, CP and DD plans in a head-to-head fashion using Student's t-test. Results: We performed 39 verification scans; each patient underwent 3 to 6 verification scan. We found no differences in doses to target or OAR structures between V and CP, V and DD, and CP and DD plans across all patients (p > 0.05). Conclusions: Automated robustness evaluation tools, CP and DD, accurately predicted dose distributions of verification (V) plans using physician-generated contours. These tools may be further developed as a potential robustness screening tool in the workflow for adaptive treatment of OPC using advanced RT techniques, reducing the need for physician-generated contours.« less

  17. Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array

    PubMed Central

    2011-01-01

    Purpose To verify the dose distribution and number of monitor units (MU) for dynamic treatment techniques like volumetric modulated single arc radiation therapy - Rapid Arc - each patient treatment plan has to be verified prior to the first treatment. The purpose of this study was to develop a patient related treatment plan verification protocol using a two dimensional ionization chamber array (MatriXX, IBA, Schwarzenbruck, Germany). Method Measurements were done to determine the dependence between response of 2D ionization chamber array, beam direction, and field size. Also the reproducibility of the measurements was checked. For the patient related verifications the original patient Rapid Arc treatment plan was projected on CT dataset of the MatriXX and the dose distribution was calculated. After irradiation of the Rapid Arc verification plans measured and calculated 2D dose distributions were compared using the gamma evaluation method implemented in the measuring software OmniPro (version 1.5, IBA, Schwarzenbruck, Germany). Results The dependence between response of 2D ionization chamber array, field size and beam direction has shown a passing rate of 99% for field sizes between 7 cm × 7 cm and 24 cm × 24 cm for measurements of single arc. For smaller and larger field sizes than 7 cm × 7 cm and 24 cm × 24 cm the passing rate was less than 99%. The reproducibility was within a passing rate of 99% and 100%. The accuracy of the whole process including the uncertainty of the measuring system, treatment planning system, linear accelerator and isocentric laser system in the treatment room was acceptable for treatment plan verification using gamma criteria of 3% and 3 mm, 2D global gamma index. Conclusion It was possible to verify the 2D dose distribution and MU of Rapid Arc treatment plans using the MatriXX. The use of the MatriXX for Rapid Arc treatment plan verification in clinical routine is reasonable. The passing rate should be 99% than the verification protocol is able to detect clinically significant errors. PMID:21342509

  18. Long-term monitoring of marine gas leakage

    NASA Astrophysics Data System (ADS)

    Spickenbom, Kai; Faber, Eckhard; Poggenburg, Jürgen; Seeger, Christian; Furche, Markus

    2010-05-01

    The sequestration of CO2 in sub-seabed geological formations is one of the Carbon Capture and Storage (CCS) strategies currently under study. Although offshore operations are significantly more expensive than comparable onshore operations, the growing public resistance against onshore CCS projects makes sub-seabed storage a promising option. Even after a thorough review of the geological setting, there is always the possibility of leakage from the reservoir. As part of the EU-financed project CO2ReMoVe (Research, Monitoring, Verification), which aims to develop innovative research and technologies for monitoring and verification of carbon dioxide geological storage, we are working on the development of submarine long-term gas flow monitoring systems. The basic design of the monitoring system builds on our experience in volcano monitoring. Early prototypes were composed of a raft floating on the surface of a mud volcano, carrying sensors for CO2 flux and concentration, data storage and transmission, and power supply by battery-buffered solar panels. The system was modified for installation in open sea by using a buoy instead of a raft and a funnel on the seafloor to collect the gas, connected by a flexible tube. This setup provides a cost-effective solution for shallow waters. However, a buoy interferes with ship traffic, and it is also difficult to adapt this design to greater water depths. These requirements can best be complied by a completely submersed system. A system for unattended long-term monitoring in a marine environment has to be extremely durable. Therefore, we focussed on developing a mechanically and electrically as simple setup as possible, which has the additional advantage of low cost. The system consists of a funnel-shaped gas collector, a sensor head and pressure housings for electronics and power supply. Since this setup is inexpensive, it can be deployed in numbers to cover larger areas. By addition of multi-channel data loggers, data transmission by acoustic modem or cable, relay stations on the seafloor or buoys etc. the infrastructure can be adapted to the environmental setting and financial budget. Prototype tests under laboratory conditions as well as field tests on natural submarine gas vents as an analogue to leaking storage sites have demonstrated the capabilities and robustness of the systems.

  19. SU-F-T-383: Robustness for Patient Setup Error in Total Body Irradiation Using Volumetric Modulated Arc Therapy (VMAT)

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Takahashi, Y; National Cancer Center, Kashiwa, Chiba; Tachibana, H

    Purpose: Total body irradiation (TBI) and total marrow irradiation (TMI) using Tomotherapy have been reported. A gantry-based linear accelerator uses one isocenter during one rotational irradiation. Thus, 3–5 isocenter points should be used for a whole plan of TBI-VMAT during smoothing out the junctional dose distribution. IGRT provides accurate and precise patient setup for the multiple junctions, however it is evident that some setup errors should occur and affect accuracy of dose distribution in the area. In this study, we evaluated the robustness for patient’s setup error in VMAT-TBI. Methods: VMAT-TBI Planning was performed in an adult whole-body human phantommore » using Eclipse. Eight full arcs with four isocenter points using 6MV-X were used to cover the entire whole body. Dose distribution was optimized using two structures of patient’s body as PTV and lung. The two arcs were shared with one isocenter and the two arcs were 5 cm-overlapped with the other two arcs. Point absolute dose using ionization-chamber and planer relative dose distribution using film in the junctional regions were performed using water-equivalent slab phantom. In the measurements, several setup errors of (+5∼−5mm) were added. Results: The result of the chamber measurement shows the deviations were within ±3% when the setup errors were within ±3 mm. In the planer evaluation, the pass ratio of gamma evaluation (3%/2mm) shows more than 90% if the errors within ±3 mm. However, there were hot/cold areas in the edge of the junction even with acceptable gamma pass ratio. 5 mm setup error caused larger hot and cold areas and the dosimetric acceptable areas were decreased in the overlapped areas. Conclusion: It can be clinically acceptable for VMAT-TBI when patient setup error is within ±3mm. Averaging effects from patient random error would be helpful to blur the hot/cold area in the junction.« less

  20. SU-E-J-44: A Novel Approach to Quantify Patient Setup and Target Motion for Real-Time Image-Guided Radiotherapy (IGRT)

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Li, S; Charpentier, P; Sayler, E

    2015-06-15

    Purpose Isocenter shifts and rotations to correct patient setup errors and organ motion cannot remedy some shape changes of large targets. We are investigating new methods in quantification of target deformation for realtime IGRT of breast and chest wall cancer. Methods Ninety-five patients of breast or chest wall cancer were accrued in an IRB-approved clinical trial of IGRT using 3D surface images acquired at daily setup and beam-on time via an in-room camera. Shifts and rotations relating to the planned reference surface were determined using iterative-closest-point alignment. Local surface displacements and target deformation are measured via a ray-surface intersection andmore » principal component analysis (PCA) of external surface, respectively. Isocenter shift, upper-abdominal displacement, and vectors of the surface projected onto the two principal components, PC1 and PC2, were evaluated for sensitivity and accuracy in detection of target deformation. Setup errors for some deformed targets were estimated by superlatively registering target volume, inner surface, or external surface in weekly CBCT or these outlines on weekly EPI. Results Setup difference according to the inner-surface, external surface, or target volume could be 1.5 cm. Video surface-guided setup agreed with EPI results to within < 0.5 cm while CBCT results were sometimes (∼20%) different from that of EPI (>0.5 cm) due to target deformation for some large breasts and some chest walls undergoing deep-breath-hold irradiation. Square root of PC1 and PC2 is very sensitive to external surface deformation and irregular breathing. Conclusion PCA of external surfaces is quick and simple way to detect target deformation in IGRT of breast and chest wall cancer. Setup corrections based on the target volume, inner surface, and external surface could be significant different. Thus, checking of target shape changes is essential for accurate image-guided patient setup and motion tracking of large deformable targets. NIH grant for the first author as cionsultant and the last author as the PI.« less

  1. 16 CFR 315.5 - Prescriber verification.

    Code of Federal Regulations, 2011 CFR

    2011-01-01

    ... time of verification request; (6) The name of a contact person at the seller's company, including... digital image of the prescription), that was presented to the seller by the patient or prescriber. (2) For...

  2. 16 CFR 315.5 - Prescriber verification.

    Code of Federal Regulations, 2010 CFR

    2010-01-01

    ... time of verification request; (6) The name of a contact person at the seller's company, including... digital image of the prescription), that was presented to the seller by the patient or prescriber. (2) For...

  3. Impact of geometric uncertainties on dose calculations for intensity modulated radiation therapy of prostate cancer

    NASA Astrophysics Data System (ADS)

    Jiang, Runqing

    Intensity-modulated radiation therapy (IMRT) uses non-uniform beam intensities within a radiation field to provide patient-specific dose shaping, resulting in a dose distribution that conforms tightly to the planning target volume (PTV). Unavoidable geometric uncertainty arising from patient repositioning and internal organ motion can lead to lower conformality index (CI) during treatment delivery, a decrease in tumor control probability (TCP) and an increase in normal tissue complication probability (NTCP). The CI of the IMRT plan depends heavily on steep dose gradients between the PTV and organ at risk (OAR). Geometric uncertainties reduce the planned dose gradients and result in a less steep or "blurred" dose gradient. The blurred dose gradients can be maximized by constraining the dose objective function in the static IMRT plan or by reducing geometric uncertainty during treatment with corrective verification imaging. Internal organ motion and setup error were evaluated simultaneously for 118 individual patients with implanted fiducials and MV electronic portal imaging (EPI). A Gaussian probability density function (PDF) is reasonable for modeling geometric uncertainties as indicated by the 118 patients group. The Gaussian PDF is patient specific and group standard deviation (SD) should not be used for accurate treatment planning for individual patients. In addition, individual SD should not be determined or predicted from small imaging samples because of random nature of the fluctuations. Frequent verification imaging should be employed in situations where geometric uncertainties are expected. Cumulative PDF data can be used for re-planning to assess accuracy of delivered dose. Group data is useful for determining worst case discrepancy between planned and delivered dose. The margins for the PTV should ideally represent true geometric uncertainties. The measured geometric uncertainties were used in this thesis to assess PTV coverage, dose to OAR, equivalent uniform dose per fraction (EUDf) and NTCP. The dose distribution including geometric uncertainties was determined from integration of the convolution of the static dose gradient with the PDF. Integration of the convolution of the static dose and derivative of the PDF can also be used to determine the dose including geometric uncertainties although this method was not investigated in detail. Local maximum dose gradient (LMDG) was determined via optimization of dose objective function by manually adjusting DVH control points or selecting beam numbers and directions during IMRT treatment planning. Minimum SD (SDmin) is used when geometric uncertainty is corrected with verification imaging. Maximum SD (SDmax) is used when the geometric uncertainty is known to be large and difficult to manage. SDmax was 4.38 mm in anterior-posterior (AP) direction, 2.70 mm in left-right (LR) direction and 4.35 mm in superior-inferior (SI) direction; SDmin was 1.1 mm in all three directions if less than 2 mm threshold was used for uncorrected fractions in every direction. EUDf is a useful QA parameter for interpreting the biological impact of geometric uncertainties on the static dose distribution. The EUD f has been used as the basis for the time-course NTCP evaluation in the thesis. Relative NTCP values are useful for comparative QA checking by normalizing known complications (e.g. reported in the RTOG studies) to specific DVH control points. For prostate cancer patients, rectal complications were evaluated from specific RTOG clinical trials and detailed evaluation of the treatment techniques (e.g. dose prescription, DVH, number of beams, bean angles). Treatment plans that did not meet DVH constraints represented additional complication risk. Geometric uncertainties improved or worsened rectal NTCP depending on individual internal organ motion within patient.

  4. Feasibility study on dosimetry verification of volumetric-modulated arc therapy-based total marrow irradiation.

    PubMed

    Liang, Yun; Kim, Gwe-Ya; Pawlicki, Todd; Mundt, Arno J; Mell, Loren K

    2013-03-04

    The purpose of this study was to develop dosimetry verification procedures for volumetric-modulated arc therapy (VMAT)-based total marrow irradiation (TMI). The VMAT based TMI plans were generated for three patients: one child and two adults. The planning target volume (PTV) was defined as bony skeleton, from head to mid-femur, with a 3 mm margin. The plan strategy similar to published studies was adopted. The PTV was divided into head and neck, chest, and pelvic regions, with separate plans each of which is composed of 2-3 arcs/fields. Multiple isocenters were evenly distributed along the patient's axial direction. The focus of this study is to establish a dosimetry quality assurance procedure involving both two-dimensional (2D) and three-dimensional (3D) volumetric verifications, which is desirable for a large PTV treated with multiple isocenters. The 2D dose verification was performed with film for gamma evaluation and absolute point dose was measured with ion chamber, with attention to the junction between neighboring plans regarding hot/cold spots. The 3D volumetric dose verification used commercial dose reconstruction software to reconstruct dose from electronic portal imaging devices (EPID) images. The gamma evaluation criteria in both 2D and 3D verification were 5% absolute point dose difference and 3 mm of distance to agreement. With film dosimetry, the overall average gamma passing rate was 98.2% and absolute dose difference was 3.9% in junction areas among the test patients; with volumetric portal dosimetry, the corresponding numbers were 90.7% and 2.4%. A dosimetry verification procedure involving both 2D and 3D was developed for VMAT-based TMI. The initial results are encouraging and warrant further investigation in clinical trials.

  5. Evaluation of setup uncertainties for single-fraction SRS by comparing two different mask-creation methods

    NASA Astrophysics Data System (ADS)

    Baek, Jong Geun; Jang, Hyun Soo; Oh, Young Kee; Lee, Hyun Jeong; Kim, Eng Chan

    2015-07-01

    The purpose of this study was to evaluate the setup uncertainties for single-fraction stereotactic radiosurgery (SF-SRS) based on clinical data with two different mask-creation methods using pretreatment con-beam computed tomography imaging guidance. Dedicated frameless fixation Brain- LAB masks for 23 patients were created as a routine mask (R-mask) making method, as explained in the BrainLAB's user manual. Alternative masks (A-masks), which were created by modifying the cover range of the R-masks for the patient's head, were used for 23 patients. The systematic errors including these for each mask and stereotactic target localizer were analyzed, and the errors were calculated as the means ± standard deviations (SD) from the left-right (LR), superior-inferior (SI), anterior-posterior (AP), and yaw setup corrections. In addition, the frequencies of the threedimensional (3D) vector length were analyzed. The values of the mean setup corrections for the R-mask in all directions were < 0.7 mm and < 0.1°, whereas the magnitudes of the SDs were relatively large compared to the mean values. In contrast, the means and SDs of the A-mask were smaller than those for the R-mask with the exception of the SD in the AP direction. The means and SDs in the yaw rotational direction for the R-mask and the A-mask system were comparable. 3D vector shifts of larger magnitude occurred more frequently for the R-mask than the A-mask. The setup uncertainties for each mask with the stereotactic localizing system had an asymmetric offset towards the positive AP direction. The A-mask-creation method, which is capable of covering the top of the patient's head, is superior to that for the R-mask, so the use of the A-mask is encouraged for SF-SRS to reduce the setup uncertainties. Moreover, careful mask-making is required to prevent possible setup uncertainties.

  6. MO-F-CAMPUS-T-05: Correct Or Not to Correct for Rotational Patient Set-Up Errors in Stereotactic Radiosurgery

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Briscoe, M; Ploquin, N; Voroney, JP

    2015-06-15

    Purpose: To quantify the effect of patient rotation in stereotactic radiation therapy and establish a threshold where rotational patient set-up errors have a significant impact on target coverage. Methods: To simulate rotational patient set-up errors, a Matlab code was created to rotate the patient dose distribution around the treatment isocentre, located centrally in the lesion, while keeping the structure contours in the original locations on the CT and MRI. Rotations of 1°, 3°, and 5° for each of the pitch, roll, and yaw, as well as simultaneous rotations of 1°, 3°, and 5° around all three axes were applied tomore » two types of brain lesions: brain metastasis and acoustic neuroma. In order to analyze multiple tumour shapes, these plans included small spherical (metastasis), elliptical (acoustic neuroma), and large irregular (metastasis) tumour structures. Dose-volume histograms and planning target volumes were compared between the planned patient positions and those with simulated rotational set-up errors. The RTOG conformity index for patient rotation was also investigated. Results: Examining the tumour volumes that received 80% of the prescription dose in the planned and rotated patient positions showed decreases in prescription dose coverage of up to 2.3%. Conformity indices for treatments with simulated rotational errors showed decreases of up to 3% compared to the original plan. For irregular lesions, degradation of 1% of the target coverage can be seen for rotations as low as 3°. Conclusions: This data shows that for elliptical or spherical targets, rotational patient set-up errors less than 3° around any or all axes do not have a significant impact on the dose delivered to the target volume or the conformity index of the plan. However the same rotational errors would have an impact on plans for irregular tumours.« less

  7. Dosimetric effects of patient rotational setup errors on prostate IMRT treatments

    NASA Astrophysics Data System (ADS)

    Fu, Weihua; Yang, Yong; Li, Xiang; Heron, Dwight E.; Saiful Huq, M.; Yue, Ning J.

    2006-10-01

    The purpose of this work is to determine dose delivery errors that could result from systematic rotational setup errors (ΔΦ) for prostate cancer patients treated with three-phase sequential boost IMRT. In order to implement this, different rotational setup errors around three Cartesian axes were simulated for five prostate patients and dosimetric indices, such as dose-volume histogram (DVH), tumour control probability (TCP), normal tissue complication probability (NTCP) and equivalent uniform dose (EUD), were employed to evaluate the corresponding dosimetric influences. Rotational setup errors were simulated by adjusting the gantry, collimator and horizontal couch angles of treatment beams and the dosimetric effects were evaluated by recomputing the dose distributions in the treatment planning system. Our results indicated that, for prostate cancer treatment with the three-phase sequential boost IMRT technique, the rotational setup errors do not have significant dosimetric impacts on the cumulative plan. Even in the worst-case scenario with ΔΦ = 3°, the prostate EUD varied within 1.5% and TCP decreased about 1%. For seminal vesicle, slightly larger influences were observed. However, EUD and TCP changes were still within 2%. The influence on sensitive structures, such as rectum and bladder, is also negligible. This study demonstrates that the rotational setup error degrades the dosimetric coverage of target volume in prostate cancer treatment to a certain degree. However, the degradation was not significant for the three-phase sequential boost prostate IMRT technique and for the margin sizes used in our institution.

  8. Reconstruction of dynamic structures of experimental setups based on measurable experimental data only

    NASA Astrophysics Data System (ADS)

    Chen, Tian-Yu; Chen, Yang; Yang, Hu-Jiang; Xiao, Jing-Hua; Hu, Gang

    2018-03-01

    Nowadays, massive amounts of data have been accumulated in various and wide fields, it has become today one of the central issues in interdisciplinary fields to analyze existing data and extract as much useful information as possible from data. It is often that the output data of systems are measurable while dynamic structures producing these data are hidden, and thus studies to reveal system structures by analyzing available data, i.e., reconstructions of systems become one of the most important tasks of information extractions. In the past, most of the works in this respect were based on theoretical analyses and numerical verifications. Direct analyses of experimental data are very rare. In physical science, most of the analyses of experimental setups were based on the first principles of physics laws, i.e., so-called top-down analyses. In this paper, we conducted an experiment of “Boer resonant instrument for forced vibration” (BRIFV) and inferred the dynamic structure of the experimental set purely from the analysis of the measurable experimental data, i.e., by applying the bottom-up strategy. Dynamics of the experimental set is strongly nonlinear and chaotic, and itʼs subjects to inevitable noises. We proposed to use high-order correlation computations to treat nonlinear dynamics; use two-time correlations to treat noise effects. By applying these approaches, we have successfully reconstructed the structure of the experimental setup, and the dynamic system reconstructed with the measured data reproduces good experimental results in a wide range of parameters.

  9. A tracking and verification system implemented in a clinical environment for partial HIPAA compliance

    NASA Astrophysics Data System (ADS)

    Guo, Bing; Documet, Jorge; Liu, Brent; King, Nelson; Shrestha, Rasu; Wang, Kevin; Huang, H. K.; Grant, Edward G.

    2006-03-01

    The paper describes the methodology for the clinical design and implementation of a Location Tracking and Verification System (LTVS) that has distinct benefits for the Imaging Department at the Healthcare Consultation Center II (HCCII), an outpatient imaging facility located on the USC Health Science Campus. A novel system for tracking and verification of patients and staff in a clinical environment using wireless and facial biometric technology to monitor and automatically identify patients and staff was developed in order to streamline patient workflow, protect against erroneous examinations and create a security zone to prevent and audit unauthorized access to patient healthcare data under the HIPAA mandate. This paper describes the system design and integration methodology based on initial clinical workflow studies within a clinical environment. An outpatient center was chosen as an initial first step for the development and implementation of this system.

  10. Assessment and quantification of patient set-up errors in nasopharyngeal cancer patients and their biological and dosimetric impact in terms of generalized equivalent uniform dose (gEUD), tumour control probability (TCP) and normal tissue complication probability (NTCP).

    PubMed

    Boughalia, A; Marcie, S; Fellah, M; Chami, S; Mekki, F

    2015-06-01

    The aim of this study is to assess and quantify patients' set-up errors using an electronic portal imaging device and to evaluate their dosimetric and biological impact in terms of generalized equivalent uniform dose (gEUD) on predictive models, such as the tumour control probability (TCP) and the normal tissue complication probability (NTCP). 20 patients treated for nasopharyngeal cancer were enrolled in the radiotherapy-oncology department of HCA. Systematic and random errors were quantified. The dosimetric and biological impact of these set-up errors on the target volume and the organ at risk (OARs) coverage were assessed using calculation of dose-volume histogram, gEUD, TCP and NTCP. For this purpose, an in-house software was developed and used. The standard deviations (1SDs) of the systematic set-up and random set-up errors were calculated for the lateral and subclavicular fields and gave the following results: ∑ = 0.63 ± (0.42) mm and σ = 3.75 ± (0.79) mm, respectively. Thus a planning organ at risk volume (PRV) margin of 3 mm was defined around the OARs, and a 5-mm margin used around the clinical target volume. The gEUD, TCP and NTCP calculations obtained with and without set-up errors have shown increased values for tumour, where ΔgEUD (tumour) = 1.94% Gy (p = 0.00721) and ΔTCP = 2.03%. The toxicity of OARs was quantified using gEUD and NTCP. The values of ΔgEUD (OARs) vary from 0.78% to 5.95% in the case of the brainstem and the optic chiasm, respectively. The corresponding ΔNTCP varies from 0.15% to 0.53%, respectively. The quantification of set-up errors has a dosimetric and biological impact on the tumour and on the OARs. The developed in-house software using the concept of gEUD, TCP and NTCP biological models has been successfully used in this study. It can be used also to optimize the treatment plan established for our patients. The gEUD, TCP and NTCP may be more suitable tools to assess the treatment plans before treating the patients.

  11. Online 3D EPID-based dose verification: Proof of concept.

    PubMed

    Spreeuw, Hanno; Rozendaal, Roel; Olaciregui-Ruiz, Igor; González, Patrick; Mans, Anton; Mijnheer, Ben; van Herk, Marcel

    2016-07-01

    Delivery errors during radiotherapy may lead to medical harm and reduced life expectancy for patients. Such serious incidents can be avoided by performing dose verification online, i.e., while the patient is being irradiated, creating the possibility of halting the linac in case of a large overdosage or underdosage. The offline EPID-based 3D in vivo dosimetry system clinically employed at our institute is in principle suited for online treatment verification, provided the system is able to complete 3D dose reconstruction and verification within 420 ms, the present acquisition time of a single EPID frame. It is the aim of this study to show that our EPID-based dosimetry system can be made fast enough to achieve online 3D in vivo dose verification. The current dose verification system was sped up in two ways. First, a new software package was developed to perform all computations that are not dependent on portal image acquisition separately, thus removing the need for doing these calculations in real time. Second, the 3D dose reconstruction algorithm was sped up via a new, multithreaded implementation. Dose verification was implemented by comparing planned with reconstructed 3D dose distributions delivered to two regions in a patient: the target volume and the nontarget volume receiving at least 10 cGy. In both volumes, the mean dose is compared, while in the nontarget volume, the near-maximum dose (D2) is compared as well. The real-time dosimetry system was tested by irradiating an anthropomorphic phantom with three VMAT plans: a 6 MV head-and-neck treatment plan, a 10 MV rectum treatment plan, and a 10 MV prostate treatment plan. In all plans, two types of serious delivery errors were introduced. The functionality of automatically halting the linac was also implemented and tested. The precomputation time per treatment was ∼180 s/treatment arc, depending on gantry angle resolution. The complete processing of a single portal frame, including dose verification, took 266 ± 11 ms on a dual octocore Intel Xeon E5-2630 CPU running at 2.40 GHz. The introduced delivery errors were detected after 5-10 s irradiation time. A prototype online 3D dose verification tool using portal imaging has been developed and successfully tested for two different kinds of gross delivery errors. Thus, online 3D dose verification has been technologically achieved.

  12. Witnessing effective entanglement over a 2 km fiber channel.

    PubMed

    Wittmann, Christoffer; Fürst, Josef; Wiechers, Carlos; Elser, Dominique; Häseler, Hauke; Lütkenhaus, Norbert; Leuchs, Gerd

    2010-03-01

    We present a fiber-based continuous-variable quantum key distribution system. In the scheme, a quantum signal of two non-orthogonal weak optical coherent states is sent through a fiber-based quantum channel. The receiver simultaneously measures conjugate quadratures of the light using two homodyne detectors. From the measured Q-function of the transmitted signal, we estimate the attenuation and the excess noise caused by the channel. The estimated excess noise originating from the channel and the channel attenuation including the quantum efficiency of the detection setup is investigated with respect to the detection of effective entanglement. The local oscillator is considered in the verification. We witness effective entanglement with a channel length of up to 2 km.

  13. Verifying the operational set-up of a radionuclide air-monitoring station.

    PubMed

    Werzi, R; Padoani, F

    2007-05-01

    A worldwide radionuclide network of 80 stations, part of the International Monitoring System, was designed to monitor compliance with the Comprehensive Nuclear-Test-Ban Treaty. After installation, the stations are certified to comply with the minimum requirements laid down by the Preparatory Commission of the Comprehensive Nuclear-Test-Ban Treaty Organization. Among the several certification tests carried out at each station, the verification of the radionuclide activity concentrations is a crucial one and is based on an independent testing of the airflow rate measurement system and of the gamma detector system, as well as on the assessment of the samples collected during parallel sampling and measured at radionuclide laboratories.

  14. Impact of radiation attenuation by a carbon fiber couch on patient dose verification

    NASA Astrophysics Data System (ADS)

    Yu, Chun-Yen; Chou, Wen-Tsae; Liao, Yi-Jen; Lee, Jeng-Hung; Liang, Ji-An; Hsu, Shih-Ming

    2017-02-01

    The aim of this study was to understand the difference between the measured and calculated irradiation attenuations obtained using two algorithms and to identify the influence of couch attenuation on patient dose verification. We performed eight tests of couch attenuation with two photon energies, two longitudinal couch positions, and two rail positions. The couch attenuation was determined using a radiation treatment planning system. The measured and calculated attenuations were compared. We also performed 12 verifications of head-and-neck and rectum cases by using a Delta phantom. The dose deviation (DD), distance to agreement (DTA), and gamma index of pencil-beam convolution (PBC) verifications were nearly the same. The agreement was least consistent for the anisotropic analytical algorithm (AAA) without the couch for the head-and-neck case, in which the DD, DTA, and gamma index were 74.4%, 99.3%, and 89%, respectively; for the rectum case, the corresponding values were 56.2%, 95.1%, and 92.4%. We suggest that dose verification should be performed using the following three metrics simultaneously: DD, DTA, and the gamma index.

  15. Surgical robot setup simulation with consistent kinematics and haptics for abdominal surgery.

    PubMed

    Hayashibe, Mitsuhiro; Suzuki, Naoki; Hattori, Asaki; Suzuki, Shigeyuki; Konishi, Kozo; Kakeji, Yoshihiro; Hashizume, Makoto

    2005-01-01

    Preoperative simulation and planning of surgical robot setup should accompany advanced robotic surgery if their advantages are to be further pursued. Feedback from the planning system will plays an essential role in computer-aided robotic surgery in addition to preoperative detailed geometric information from patient CT/MRI images. Surgical robot setup simulation systems for appropriate trocar site placement have been developed especially for abdominal surgery. The motion of the surgical robot can be simulated and rehearsed with kinematic constraints at the trocar site, and the inverse-kinematics of the robot. Results from simulation using clinical patient data verify the effectiveness of the proposed system.

  16. Acoustic Noise Levels of Dental Equipments and Its Association with Fear and Annoyance Levels among Patients Attending Different Dental Clinic Setups in Jaipur, India

    PubMed Central

    Ganta, Shravani; Nagaraj, Anup; Pareek, Sonia; Atri, Mansi; Singh, Kushpal; Sidiq, Mohsin

    2014-01-01

    Background: Noise is a source of pervasive occupational hazard for practicing dentists and the patients. The sources of dental sounds by various dental equipments can pose as a potential hazard to hearing system and add to the annoyance levels of the patients. The aim of the study was to analyze the noise levels from various equipments and evaluate the effect of acoustic noise stimulus on dental fear and annoyance levels among patients attending different dental clinic setups in Jaipur, India. Methodology: The sampling frame comprised of 180 patients, which included 90 patients attending 10 different private clinics and 90 patients attending a Dental College in Jaipur. The levels of Acoustic Noise Stimulus originating from different equipments were determined using a precision sound level meter/decibulometer. Dental fear among patients was measured using Dental Fear Scale (DFS). Results: Statistical analysis was performed using chi square test and unpaired t-test. The mean background noise levels were found to be maximum in the pre-clinical setup/ laboratory areas (69.23+2.20). Females and the patients attending dental college setup encountered more fear on seeing the drill as compared to the patients attending private clinics (p<0.001). Conclusion: The sources of dental sounds can pose as a potential hazard to hearing system. It was analyzed that the environment in the clinics can directly have an effect on the fear and annoyance levels of patients. Hence it is necessary control the noise from various dental equipments to reduce the fear of patients from visiting a dental clinic. PMID:24959512

  17. Local Setup Reproducibility of the Spinal Column When Using Intensity-Modulated Radiation Therapy for Craniospinal Irradiation With Patient in Supine Position

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Stoiber, Eva Maria, E-mail: eva.stoiber@med.uni-heidelberg.de; Department of Medical Physics, German Cancer Research Center, Heidelberg; Giske, Kristina

    Purpose: To evaluate local positioning errors of the lumbar spine during fractionated intensity-modulated radiotherapy of patients treated with craniospinal irradiation and to assess the impact of rotational error correction on these uncertainties for one patient setup correction strategy. Methods and Materials: 8 patients (6 adults, 2 children) treated with helical tomotherapy for craniospinal irradiation were retrospectively chosen for this analysis. Patients were immobilized with a deep-drawn Aquaplast head mask. Additionally to daily megavoltage control computed tomography scans of the skull, once-a-week positioning of the lumbar spine was assessed. Therefore, patient setup was corrected by a target point correction, derived frommore » a registration of the patient's skull. The residual positioning variations of the lumbar spine were evaluated applying a rigid-registration algorithm. The impact of different rotational error corrections was simulated. Results: After target point correction, residual local positioning errors of the lumbar spine varied considerably. Craniocaudal axis rotational error correction did not improve or deteriorate these translational errors, whereas simulation of a rotational error correction of the right-left and anterior-posterior axis increased these errors by a factor of 2 to 3. Conclusion: The patient fixation used allows for deformations between the patient's skull and spine. Therefore, for the setup correction strategy evaluated in this study, generous margins for the lumbar spinal target volume are needed to prevent a local geographic miss. With any applied correction strategy, it needs to be evaluated whether or not a rotational error correction is beneficial.« less

  18. Automatic detection of patient identification and positioning errors in radiation therapy treatment using 3-dimensional setup images.

    PubMed

    Jani, Shyam S; Low, Daniel A; Lamb, James M

    2015-01-01

    To develop an automated system that detects patient identification and positioning errors between 3-dimensional computed tomography (CT) and kilovoltage CT planning images. Planning kilovoltage CT images were collected for head and neck (H&N), pelvis, and spine treatments with corresponding 3-dimensional cone beam CT and megavoltage CT setup images from TrueBeam and TomoTherapy units, respectively. Patient identification errors were simulated by registering setup and planning images from different patients. For positioning errors, setup and planning images were misaligned by 1 to 5 cm in the 6 anatomical directions for H&N and pelvis patients. Spinal misalignments were simulated by misaligning to adjacent vertebral bodies. Image pairs were assessed using commonly used image similarity metrics as well as custom-designed metrics. Linear discriminant analysis classification models were trained and tested on the imaging datasets, and misclassification error (MCE), sensitivity, and specificity parameters were estimated using 10-fold cross-validation. For patient identification, our workflow produced MCE estimates of 0.66%, 1.67%, and 0% for H&N, pelvis, and spine TomoTherapy images, respectively. Sensitivity and specificity ranged from 97.5% to 100%. MCEs of 3.5%, 2.3%, and 2.1% were obtained for TrueBeam images of the above sites, respectively, with sensitivity and specificity estimates between 95.4% and 97.7%. MCEs for 1-cm H&N/pelvis misalignments were 1.3%/5.1% and 9.1%/8.6% for TomoTherapy and TrueBeam images, respectively. Two-centimeter MCE estimates were 0.4%/1.6% and 3.1/3.2%, respectively. MCEs for vertebral body misalignments were 4.8% and 3.6% for TomoTherapy and TrueBeam images, respectively. Patient identification and gross misalignment errors can be robustly and automatically detected using 3-dimensional setup images of different energies across 3 commonly treated anatomical sites. Copyright © 2015 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.

  19. SU-F-T-308: Mobius FX Evaluation and Comparison Against a Commercial 4D Detector Array for VMAT Plan QA

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Vazquez Quino, L; Huerta Hernandez, C; Morrow, A

    2016-06-15

    Purpose: To evaluate the use of MobiusFX as a pre-treatment verification IMRT QA tool and compare it with a commercial 4D detector array for VMAT plan QA. Methods: 15 VMAT plan QA of different treatment sites were delivered and measured by traditional means with the 4D detector array ArcCheck (Sun Nuclear corporation) and at the same time measurement in linac treatment logs (Varian Dynalogs files) were analyzed from the same delivery with MobiusFX software (Mobius Medical Systems). VMAT plan QAs created in Eclipse treatment planning system (Varian) in a TrueBeam linac machine (Varian) were delivered and analyzed with the gammamore » analysis routine from SNPA software (Sun Nuclear corporation). Results: Comparable results in terms of the gamma analysis with 99.06% average gamma passing with 3%,3mm passing rate is observed in the comparison among MobiusFX, ArcCheck measurements, and the Treatment Planning System dose calculated. When going to a stricter criterion (1%,1mm) larger discrepancies are observed in different regions of the measurements with an average gamma of 66.24% between MobiusFX and ArcCheck. Conclusion: This work indicates the potential for using MobiusFX as a routine pre-treatment patient specific IMRT method for quality assurance purposes and its advantages as a phantom-less method which reduce the time for IMRT QA measurement. MobiusFX is capable of produce similar results of those by traditional methods used for patient specific pre-treatment verification VMAT QA. Even the gamma results comparing to the TPS are similar the analysis of both methods show that the errors being identified by each method are found in different regions. Traditional methods like ArcCheck are sensitive to setup errors and dose difference errors coming from the linac output. On the other hand linac log files analysis record different errors in the VMAT QA associated with the MLCs and gantry motion that by traditional methods cannot be detected.« less

  20. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Moteabbed, M; Trofimov, A; Sharp, G

    Purpose: To investigate the impact of anatomy/setup variations on standard vs. hypofractionated anterolateral pencil beam scanning (PBS) proton therapy for prostate cancer. Methods: Six prostate cancer patients treated with double-scattering proton therapy, who underwent weekly verification CT scans were selected. Implanted fiducials were used for localization, and endorectal balloons for immobilization. New PBS plans using combination of lateral and anterior-oblique (AO) (±35 deg) beams were created. AO beams were added to spare the femoral heads during hypofractionation. Lateral beams delivered 50.4 Gy(RBE) to prostate plus 5-15mm of seminal vesicles and AO beams 28.8 Gy(RBE) to prostate, in 44 fractions. PTVmore » was laterally expanded by 2.5% to account for range uncertainty. No range margins were applied for AO beams, assuming delivery with in-vivo range verification. Field-specific apertures with 1.2cm margin were used. Spot size was ∼9.5mm sigma for 172MeV @isocenter in air. Plans were optimized as single-field-uniform-dose with ∼5% maximum non-uniformity. The planned dose was recomputed on each weekly CT after aligning the fiducials with the simulation CT, scaled and accumulated via deformable image registration. Hypofractionated treatments with 12 and 5 fractions were considered. Equivalent doses were calculated for prostate (α/β= 1.5Gy), bladder and rectum (α/β= 3Gy). Results: The biological equivalent prostate dose was 86.2 and 92.9 Gyeq for the hypofractionation scenarios at 4.32 and 7.35 Gy/fx, respectively. The equivalent prostate D98 was degraded by on average 2.7 Gyeq for standard, and 3.1 and 4.0 Gyeq for the hypofractionated plans after accumulation. Differences between accumulated and planned Dmean/D2/EUD were generally reduced when reducing the number of fractions for bladder and rectum. The average Dmean/D2/EUD differences over all patients and organs-at-risk were 0.74/4.0/9.23, 0.49/3.64/5.51, 0.37/3.21/3.49 Gyeq for 44, 12 and 5 fractions. Conclusion: Hypofractionation makes proton therapy of prostate more susceptible to interfractional motion-induced target dose degradation compared to the standard fractionation.« less

  1. MO-FG-CAMPUS-TeP1-04: Pseudo-In-Vivo Dose Verification of a New Mono-Isocentric Technique for the Treatment of Multiple Brain Metastases

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Pappas, E P; Makris, D; Lahanas, V

    2016-06-15

    Purpose: To validate dose calculation and delivery accuracy of a recently introduced mono-isocentric technique for the treatment of multiple brain metastases in a realistic clinical case. Methods: Anonymized CT scans of a patient were used to model a hollow phantom that duplicates anatomy of the skull. A 3D printer was used to construct the phantom of a radiologically bone-equivalent material. The hollow phantom was subsequently filled with a polymer gel 3D dosimeter which also acted as a water-equivalent material. Irradiation plan consisted of 5 targets and was identical to the one delivered to the specific patient except for the prescriptionmore » dose which was optimized to match the gel dose-response characteristics. Dose delivery was performed using a single setup isocenter dynamic conformal arcs technique. Gel dose read-out was carried out by a 1.5 T MRI scanner. All steps of the corresponding patient’s treatment protocol were strictly followed providing an end-to-end quality assurance test. Pseudo-in-vivo measured 3D dose distribution and calculated one were compared in terms of spatial agreement, dose profiles, 3D gamma indices (5%/2mm, 20% dose threshold), DVHs and DVH metrics. Results: MR-identified polymerized areas and calculated high dose regions were found to agree within 1.5 mm for all targets, taking into account all sources of spatial uncertainties involved (i.e., set-up errors, MR-related geometric distortions and registration inaccuracies). Good dosimetric agreement was observed in the vast majority of the examined profiles. 3D gamma index passing rate reached 91%. DVH and corresponding metrics comparison resulted in a satisfying agreement between measured and calculated datasets within targets and selected organs-at-risk. Conclusion: A novel, pseudo-in-vivo QA test was implemented to validate spatial and dosimetric accuracy in treatment of multiple metastases. End-to-end testing demonstrated that our gel dosimetry phantom is suited for such QA procedures, allowing for 3D analysis of both targeting placement and dose.« less

  2. SU-C-BRD-02: A Team Focused Clinical Implementation and Failure Mode and Effects Analysis of HDR Skin Brachytherapy Using Valencia and Leipzig Surface Applicators

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sayler, E; Harrison, A; Eldredge-Hindy, H

    Purpose: and Leipzig applicators (VLAs) are single-channel brachytherapy surface applicators used to treat skin lesions up to 2cm diameter. Source dwell times can be calculated and entered manually after clinical set-up or ultrasound. This procedure differs dramatically from CT-based planning; the novelty and unfamiliarity could lead to severe errors. To build layers of safety and ensure quality, a multidisciplinary team created a protocol and applied Failure Modes and Effects Analysis (FMEA) to the clinical procedure for HDR VLA skin treatments. Methods: team including physicists, physicians, nurses, therapists, residents, and administration developed a clinical procedure for VLA treatment. The procedure wasmore » evaluated using FMEA. Failure modes were identified and scored by severity, occurrence, and detection. The clinical procedure was revised to address high-scoring process nodes. Results: Several key components were added to the clinical procedure to minimize risk probability numbers (RPN): -Treatments are reviewed at weekly QA rounds, where physicians discuss diagnosis, prescription, applicator selection, and set-up. Peer review reduces the likelihood of an inappropriate treatment regime. -A template for HDR skin treatments was established in the clinical EMR system to standardize treatment instructions. This reduces the chances of miscommunication between the physician and planning physicist, and increases the detectability of an error during the physics second check. -A screen check was implemented during the second check to increase detectability of an error. -To reduce error probability, the treatment plan worksheet was designed to display plan parameters in a format visually similar to the treatment console display. This facilitates data entry and verification. -VLAs are color-coded and labeled to match the EMR prescriptions, which simplifies in-room selection and verification. Conclusion: Multidisciplinary planning and FMEA increased delectability and reduced error probability during VLA HDR Brachytherapy. This clinical model may be useful to institutions implementing similar procedures.« less

  3. Verification bias: an under-recognized source of error in assessing the efficacy of MRI of the meniscii.

    PubMed

    Richardson, Michael L; Petscavage, Jonelle M

    2011-11-01

    The sensitivity and specificity of magnetic resonance imaging (MRI) for diagnosis of meniscal tears has been studied extensively, with tears usually verified by surgery. However, surgically unverified cases are often not considered in these studies, leading to verification bias, which can falsely increase the sensitivity and decrease the specificity estimates. Our study suggests that such bias may be very common in the meniscal MRI literature, and illustrates techniques to detect and correct for such bias. PubMed was searched for articles estimating sensitivity and specificity of MRI for meniscal tears. These were assessed for verification bias, deemed potentially present if a study included any patients whose MRI findings were not surgically verified. Retrospective global sensitivity analysis (GSA) was performed when possible. Thirty-nine of the 314 studies retrieved from PubMed specifically dealt with meniscal tears. All 39 included unverified patients, and hence, potential verification bias. Only seven articles included sufficient information to perform GSA. Of these, one showed definite verification bias, two showed no bias, and four others showed bias within certain ranges of disease prevalence. Only 9 of 39 acknowledged the possibility of verification bias. Verification bias is underrecognized and potentially common in published estimates of the sensitivity and specificity of MRI for the diagnosis of meniscal tears. When possible, it should be avoided by proper study design. If unavoidable, it should be acknowledged. Investigators should tabulate unverified as well as verified data. Finally, verification bias should be estimated; if present, corrected estimates of sensitivity and specificity should be used. Our online web-based calculator makes this process relatively easy. Copyright © 2011 AUR. Published by Elsevier Inc. All rights reserved.

  4. Comparison of setup accuracy of three different image assessment methods for tangential breast radiotherapy.

    PubMed

    Batumalai, Vikneswary; Phan, Penny; Choong, Callie; Holloway, Lois; Delaney, Geoff P

    2016-12-01

    To compare the differences in setup errors measured with electronic portal image (EPI) and cone-beam computed tomography (CBCT) in patients undergoing tangential breast radiotherapy (RT). Relationship between setup errors, body mass index (BMI) and breast size was assessed. Twenty-five patients undergoing postoperative RT to the breast were consented for this study. Weekly CBCT scans were acquired and retrospectively registered to the planning CT in three dimensions, first using bony anatomy for bony registration (CBCT-B) and again using breast tissue outline for soft tissue registration (CBCT-S). Digitally reconstructed radiographs (DRR) generated from CBCT to simulate EPI were compared to the planning DRR using bony anatomy in the V (parallel to the cranio-caudal axis) and U (perpendicular to V) planes. The systematic (Σ) and random (σ) errors were calculated and correlated with BMI and breast size. The systematic and random errors for EPI (Σ V = 3.7 mm, Σ U = 2.8 mm and σ V = 2.9 mm, σ U = 2.5) and CBCT-B (Σ V = 3.5 mm, Σ U = 3.4 mm and σ V = 2.8 mm, σ U = 2.8) were of similar magnitude in the V and U planes. Similarly, the differences in setup errors for CBCT-B and CBCT-S in three dimensions were less than 1 mm. Only CBCT-S setup error correlated with BMI and breast size. CBCT and EPI show insignificant variation in their ability to detect setup error. These findings suggest no significant differences that would make one modality considered superior over the other and EPI should remain the standard of care for most patients. However, there is a correlation with breast size, BMI and setup error as detected by CBCT-S, justifying the use of CBCT-S for larger patients. © 2016 The Authors. Journal of Medical Radiation Sciences published by John Wiley & Sons Australia, Ltd on behalf of Australian Society of Medical Imaging and Radiation Therapy and New Zealand Institute of Medical Radiation Technology.

  5. Usefulness of biological fingerprint in magnetic resonance imaging for patient verification.

    PubMed

    Ueda, Yasuyuki; Morishita, Junji; Kudomi, Shohei; Ueda, Katsuhiko

    2016-09-01

    The purpose of our study is to investigate the feasibility of automated patient verification using multi-planar reconstruction (MPR) images generated from three-dimensional magnetic resonance (MR) imaging of the brain. Several anatomy-related MPR images generated from three-dimensional fast scout scan of each MR examination were used as biological fingerprint images in this study. The database of this study consisted of 730 temporal pairs of MR examination of the brain. We calculated the correlation value between current and prior biological fingerprint images of the same patient and also all combinations of two images for different patients to evaluate the effectiveness of our method for patient verification. The best performance of our system were as follows: a half-total error rate of 1.59 % with a false acceptance rate of 0.023 % and a false rejection rate of 3.15 %, an equal error rate of 1.37 %, and a rank-one identification rate of 98.6 %. Our method makes it possible to verify the identity of the patient using only some existing medical images without the addition of incidental equipment. Also, our method will contribute to patient misidentification error management caused by human errors.

  6. In vivo dose verification of IMRT treated head and neck cancer patients.

    PubMed

    Engström, Per E; Haraldsson, Pia; Landberg, Torsten; Sand Hansen, Hanne; Aage Engelholm, Svend; Nyström, Håkan

    2005-01-01

    An independent in vivo dose verification procedure for IMRT treatments of head and neck cancers was developed. Results of 177 intracavitary TLD measurements from 10 patients are presented. The study includes data from 10 patients with cancer of the rhinopharynx or the thyroid treated with dynamic IMRT. Dose verification was performed by insertion of a flexible naso-oesophageal tube containing TLD rods and markers for EPID and simulator image detection. Part of the study focussed on investigating the accuracy of the TPS calculations in the presence of inhomogeneities. Phantom measurements and Monte Carlo simulations were performed for a number of geometries involving lateral electronic disequilibrium and steep density shifts. The in vivo TLD measurements correlated well with the predictions of the treatment planning system with a measured/calculated dose ratio of 1.002+/-0.051 (1 SD, N=177). The measurements were easily performed and well tolerated by the patients. We conclude that in vivo intracavitary dosimetry with TLD is suitable and accurate for dose determination in intensity-modulated beams.

  7. Assessment and quantification of patient set-up errors in nasopharyngeal cancer patients and their biological and dosimetric impact in terms of generalized equivalent uniform dose (gEUD), tumour control probability (TCP) and normal tissue complication probability (NTCP)

    PubMed Central

    Marcie, S; Fellah, M; Chami, S; Mekki, F

    2015-01-01

    Objective: The aim of this study is to assess and quantify patients' set-up errors using an electronic portal imaging device and to evaluate their dosimetric and biological impact in terms of generalized equivalent uniform dose (gEUD) on predictive models, such as the tumour control probability (TCP) and the normal tissue complication probability (NTCP). Methods: 20 patients treated for nasopharyngeal cancer were enrolled in the radiotherapy–oncology department of HCA. Systematic and random errors were quantified. The dosimetric and biological impact of these set-up errors on the target volume and the organ at risk (OARs) coverage were assessed using calculation of dose–volume histogram, gEUD, TCP and NTCP. For this purpose, an in-house software was developed and used. Results: The standard deviations (1SDs) of the systematic set-up and random set-up errors were calculated for the lateral and subclavicular fields and gave the following results: ∑ = 0.63 ± (0.42) mm and σ = 3.75 ± (0.79) mm, respectively. Thus a planning organ at risk volume (PRV) margin of 3 mm was defined around the OARs, and a 5-mm margin used around the clinical target volume. The gEUD, TCP and NTCP calculations obtained with and without set-up errors have shown increased values for tumour, where ΔgEUD (tumour) = 1.94% Gy (p = 0.00721) and ΔTCP = 2.03%. The toxicity of OARs was quantified using gEUD and NTCP. The values of ΔgEUD (OARs) vary from 0.78% to 5.95% in the case of the brainstem and the optic chiasm, respectively. The corresponding ΔNTCP varies from 0.15% to 0.53%, respectively. Conclusion: The quantification of set-up errors has a dosimetric and biological impact on the tumour and on the OARs. The developed in-house software using the concept of gEUD, TCP and NTCP biological models has been successfully used in this study. It can be used also to optimize the treatment plan established for our patients. Advances in knowledge: The gEUD, TCP and NTCP may be more suitable tools to assess the treatment plans before treating the patients. PMID:25882689

  8. Online 3D EPID-based dose verification: Proof of concept

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Spreeuw, Hanno; Rozendaal, Roel, E-mail: r.rozenda

    Purpose: Delivery errors during radiotherapy may lead to medical harm and reduced life expectancy for patients. Such serious incidents can be avoided by performing dose verification online, i.e., while the patient is being irradiated, creating the possibility of halting the linac in case of a large overdosage or underdosage. The offline EPID-based 3D in vivo dosimetry system clinically employed at our institute is in principle suited for online treatment verification, provided the system is able to complete 3D dose reconstruction and verification within 420 ms, the present acquisition time of a single EPID frame. It is the aim of thismore » study to show that our EPID-based dosimetry system can be made fast enough to achieve online 3D in vivo dose verification. Methods: The current dose verification system was sped up in two ways. First, a new software package was developed to perform all computations that are not dependent on portal image acquisition separately, thus removing the need for doing these calculations in real time. Second, the 3D dose reconstruction algorithm was sped up via a new, multithreaded implementation. Dose verification was implemented by comparing planned with reconstructed 3D dose distributions delivered to two regions in a patient: the target volume and the nontarget volume receiving at least 10 cGy. In both volumes, the mean dose is compared, while in the nontarget volume, the near-maximum dose (D2) is compared as well. The real-time dosimetry system was tested by irradiating an anthropomorphic phantom with three VMAT plans: a 6 MV head-and-neck treatment plan, a 10 MV rectum treatment plan, and a 10 MV prostate treatment plan. In all plans, two types of serious delivery errors were introduced. The functionality of automatically halting the linac was also implemented and tested. Results: The precomputation time per treatment was ∼180 s/treatment arc, depending on gantry angle resolution. The complete processing of a single portal frame, including dose verification, took 266 ± 11 ms on a dual octocore Intel Xeon E5-2630 CPU running at 2.40 GHz. The introduced delivery errors were detected after 5–10 s irradiation time. Conclusions: A prototype online 3D dose verification tool using portal imaging has been developed and successfully tested for two different kinds of gross delivery errors. Thus, online 3D dose verification has been technologically achieved.« less

  9. A quantification of the effectiveness of EPID dosimetry and software-based plan verification systems in detecting incidents in radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Bojechko, Casey; Phillps, Mark; Kalet, Alan

    Purpose: Complex treatments in radiation therapy require robust verification in order to prevent errors that can adversely affect the patient. For this purpose, the authors estimate the effectiveness of detecting errors with a “defense in depth” system composed of electronic portal imaging device (EPID) based dosimetry and a software-based system composed of rules-based and Bayesian network verifications. Methods: The authors analyzed incidents with a high potential severity score, scored as a 3 or 4 on a 4 point scale, recorded in an in-house voluntary incident reporting system, collected from February 2012 to August 2014. The incidents were categorized into differentmore » failure modes. The detectability, defined as the number of incidents that are detectable divided total number of incidents, was calculated for each failure mode. Results: In total, 343 incidents were used in this study. Of the incidents 67% were related to photon external beam therapy (EBRT). The majority of the EBRT incidents were related to patient positioning and only a small number of these could be detected by EPID dosimetry when performed prior to treatment (6%). A large fraction could be detected by in vivo dosimetry performed during the first fraction (74%). Rules-based and Bayesian network verifications were found to be complimentary to EPID dosimetry, able to detect errors related to patient prescriptions and documentation, and errors unrelated to photon EBRT. Combining all of the verification steps together, 91% of all EBRT incidents could be detected. Conclusions: This study shows that the defense in depth system is potentially able to detect a large majority of incidents. The most effective EPID-based dosimetry verification is in vivo measurements during the first fraction and is complemented by rules-based and Bayesian network plan checking.« less

  10. Microionization chamber for reference dosimetry in IMRT verification: clinical implications on OAR dosimetric errors

    NASA Astrophysics Data System (ADS)

    Sánchez-Doblado, Francisco; Capote, Roberto; Leal, Antonio; Roselló, Joan V.; Lagares, Juan I.; Arráns, Rafael; Hartmann, Günther H.

    2005-03-01

    Intensity modulated radiotherapy (IMRT) has become a treatment of choice in many oncological institutions. Small fields or beamlets with sizes of 1 to 5 cm2 are now routinely used in IMRT delivery. Therefore small ionization chambers (IC) with sensitive volumes <=0.1 cm3are generally used for dose verification of an IMRT treatment. The measurement conditions during verification may be quite different from reference conditions normally encountered in clinical beam calibration, so dosimetry of these narrow photon beams pertains to the so-called non-reference conditions for beam calibration. This work aims at estimating the error made when measuring the organ at risk's (OAR) absolute dose by a micro ion chamber (μIC) in a typical IMRT treatment. The dose error comes from the assumption that the dosimetric parameters determining the absolute dose are the same as for the reference conditions. We have selected two clinical cases, treated by IMRT, for our dose error evaluations. Detailed geometrical simulation of the μIC and the dose verification set-up was performed. The Monte Carlo (MC) simulation allows us to calculate the dose measured by the chamber as a dose averaged over the air cavity within the ion-chamber active volume (Dair). The absorbed dose to water (Dwater) is derived as the dose deposited inside the same volume, in the same geometrical position, filled and surrounded by water in the absence of the ion chamber. Therefore, the Dwater/Dair dose ratio is the MC estimator of the total correction factor needed to convert the absorbed dose in air into the absorbed dose in water. The dose ratio was calculated for the μIC located at the isocentre within the OARs for both clinical cases. The clinical impact of the calculated dose error was found to be negligible for the studied IMRT treatments.

  11. Determining the hospital trauma financial impact in a statewide trauma system.

    PubMed

    Mabry, Charles D; Kalkwarf, Kyle J; Betzold, Richard D; Spencer, Horace J; Robertson, Ronald D; Sutherland, Michael J; Maxson, Robert T

    2015-04-01

    There have been no comprehensive studies across an organized statewide trauma system using a standardized method to determine cost. Trauma financial impact includes the following costs: verification, response, and patient care cost (PCC). We conducted a survey of participating trauma centers (TCs) for federal fiscal year 2012, including separate accounting for verification and response costs. Patient care cost was merged with their trauma registry data. Seventy-five percent of the 2012 state trauma registry had data submitted. Each TC's reasonable cost from the Medicare Cost Report was adjusted to remove embedded costs for response and verification. Cost-to-charge ratios were used to give uniform PCC across the state. Median (mean ± SD) costs per patient for TC response and verification for Level I and II centers were $1,689 ($1,492 ± $647) and $450 ($636 ± $431) for Level III and IV centers. Patient care cost-median (mean ± SD) costs for patients with a length of stay >2 days rose with increasing Injury Severity Score (ISS): ISS <9: $6,787 ($8,827 ± $8,165), ISS 9 to 15: $10,390 ($14,340 ± $18,395); ISS 16 to 25: $15,698 ($23,615 ± $21,883); and ISS 25+: $29,792 ($41,407 ± $41,621), and with higher level of TC: Level I: $13,712 ($23,241 ± $29,164); Level II: $8,555 ($13,515 ± $15,296); and Levels III and IV: $8,115 ($10,719 ± $11,827). Patient care cost rose with increasing ISS, length of stay, ICU days, and ventilator days for patients with length of stay >2 days and ISS 9+. Level I centers had the highest mean ISS, length of stay, ICU days, and ventilator days, along with the highest PCC. Lesser trauma accounted for lower charges, payments, and PCC for Level II, III, and IV TCs, and the margin was variable. Verification and response costs per patient were highest for Level I and II TCs. Copyright © 2015 American College of Surgeons. Published by Elsevier Inc. All rights reserved.

  12. Ensemble sea ice forecast for predicting compressive situations in the Baltic Sea

    NASA Astrophysics Data System (ADS)

    Lehtiranta, Jonni; Lensu, Mikko; Kokkonen, Iiro; Haapala, Jari

    2017-04-01

    Forecasting of sea ice hazards is important for winter shipping in the Baltic Sea. In current numerical models the ice thickness distribution and drift are captured well, but compressive situations are often missing from forecast products. Its inclusion is requested by the shipping community, as compression poses a threat to ship operations. As compressing ice is capable of stopping ships for days and even damaging them, its inclusion in ice forecasts is vital. However, we have found that compression can not be predicted well in a deterministic forecast, since it can be a local and a quickly changing phenomenon. It is also very sensitive to small changes in the wind speed and direction, the prevailing ice conditions, and the model parameters. Thus, a probabilistic ensemble simulation is needed to produce a meaningful compression forecast. An ensemble model setup was developed in the SafeWIN project for this purpose. It uses the HELMI multicategory ice model, which was amended for making simulations in parallel. The ensemble was built by perturbing the atmospheric forcing and the physical parameters of the ice pack. The model setup will provide probabilistic forecasts for the compression in the Baltic sea ice. Additionally the model setup provides insight into the uncertainties related to different model parameters and their impact on the model results. We have completed several hindcast simulations for the Baltic Sea for verification purposes. These results are shown to match compression reports gathered from ships. In addition, an ensemble forecast is in preoperational testing phase and its first evaluation will be presented in this work.

  13. Sci-Thur PM – Brachytherapy 05: Surface Collimation Applied to Superficial Flap High Dose-Rate Brachytherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Liu, Derek; Sabondjian, Eric; Lawrence, Kailin

    Purpose: To apply surface collimation for superficial flap HDR skin brachytherapy utilizing common clinical resources and to demonstrate the potential for OAR dose reduction within a clinically relevant setting. Methods: Two phantom setups were used. 3 mm lead collimation was applied to a solid slab phantom to determine appropriate geometries relating to collimation and dwell activation. The same collimation was applied to the temple of an anthropomorphic head phantom to demonstrate lens dose reduction. Each setup was simulated and planned to deliver 400 cGy to a 3 cm circular target to 3 mm depth. The control and collimated irradiations weremore » sequentially measured using calibrated radiochromic films. Results: Collimation for the slab phantom attenuated the dose beyond the collimator opening, decreasing the fall-off distances by half and reducing the area of healthy skin irradiated. Target coverage can be negatively impacted by a tight collimation margin, with the required margin approximated by the primary beam geometric penumbra. Surface collimation applied to the head phantom similarly attenuated the surrounding normal tissue dose while reducing the lens dose from 84 to 68 cGy. To ensure consistent setup between simulation and treatment, additional QA was performed including collimator markup, accounting for collimator placement uncertainties, standoff distance verification, and in vivo dosimetry. Conclusions: Surface collimation was shown to reduce normal tissue dose without compromising target coverage. Lens dose reduction was demonstrated on an anthropomorphic phantom within a clinical setting. Additional QA is proposed to ensure treatment fidelity.« less

  14. SU-E-J-37: Feasibility of Utilizing Carbon Fiducials to Increase Localization Accuracy of Lumpectomy Cavity for Partial Breast Irradiation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zhang, Y; Hieken, T; Mutter, R

    2015-06-15

    Purpose To investigate the feasibility of utilizing carbon fiducials to increase localization accuracy of lumpectomy cavity for partial breast irradiation (PBI). Methods Carbon fiducials were placed intraoperatively in the lumpectomy cavity following resection of breast cancer in 11 patients. The patients were scheduled to receive whole breast irradiation (WBI) with a boost or 3D-conformal PBI. WBI patients were initially setup to skin tattoos using lasers, followed by orthogonal kV on-board-imaging (OBI) matching to bone per clinical practice. Cone beam CT (CBCT) was acquired weekly for offline review. For the boost component of WBI and PBI, patients were setup with lasers,more » followed by OBI matching to fiducials, with final alignment by CBCT matching to fiducials. Using carbon fiducials as a surrogate for the lumpectomy cavity and CBCT matching to fiducials as the gold standard, setup uncertainties to lasers, OBI bone, OBI fiducials, and CBCT breast were compared. Results Minimal imaging artifacts were introduced by fiducials on the planning CT and CBCT. The fiducials were sufficiently visible on OBI for online localization. The mean magnitude and standard deviation of setup errors were 8.4mm ± 5.3 mm (n=84), 7.3mm ± 3.7mm (n=87), 2.2mm ± 1.6mm (n=40) and 4.8mm ± 2.6mm (n=87), for lasers, OBI bone, OBI fiducials and CBCT breast tissue, respectively. Significant migration occurred in one of 39 implanted fiducials in a patient with a large postoperative seroma. Conclusion OBI carbon fiducial-based setup can improve localization accuracy with minimal imaging artifacts. With increased localization accuracy, setup uncertainties can be reduced from 8mm using OBI bone matching to 3mm using OBI fiducial matching for PBI treatment. This work demonstrates the feasibility of utilizing carbon fiducials to increase localization accuracy to the lumpectomy cavity for PBI. This may be particularly attractive for localization in the setting of proton therapy and other scenarios in which metal clips are contraindicated.« less

  15. Optimization and Verification of a Brushless DC-Motor for Cryogenic Mechanisms

    NASA Astrophysics Data System (ADS)

    Eggens, M.; van Loon, D.; Smit, H. P.; Jellema, W.; Dieleman, P.; Detrain, A.; Stokroos, M.; Nieuwenhuizen, A. C. T.

    2013-09-01

    In this paper we report on the results of the investigation on the feasibility of a cryogenic motor for a Filter Wheel Mechanism (FWM) for the instrument SpicA FAR-infrared Instrument (SAFARI). The maximum allowed dissipation of 1 mW is a key requirement, as a result of the limited cooling resources of the satellite. Therefore a quasi 3D electromagnetic (EM) model of a Brushless DC (BLDC) motor has been developed. To withstand the severe launch loads a mechanical concept has been designed to limit the friction torque in the bearings. The model was verified by room temperature and cryogenic measurements on an existing motor from the test setup. The model shows that the proposed BLDC motor design fulfills the requirements.

  16. Simulation and Analysis of Converging Shock Wave Test Problems

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ramsey, Scott D.; Shashkov, Mikhail J.

    2012-06-21

    Results and analysis pertaining to the simulation of the Guderley converging shock wave test problem (and associated code verification hydrodynamics test problems involving converging shock waves) in the LANL ASC radiation-hydrodynamics code xRAGE are presented. One-dimensional (1D) spherical and two-dimensional (2D) axi-symmetric geometric setups are utilized and evaluated in this study, as is an instantiation of the xRAGE adaptive mesh refinement capability. For the 2D simulations, a 'Surrogate Guderley' test problem is developed and used to obviate subtleties inherent to the true Guderley solution's initialization on a square grid, while still maintaining a high degree of fidelity to the originalmore » problem, and minimally straining the general credibility of associated analysis and conclusions.« less

  17. Communication Satellite Payload Special Check out Equipment (SCOE) for Satellite Testing

    NASA Astrophysics Data System (ADS)

    Subhani, Noman

    2016-07-01

    This paper presents Payload Special Check out Equipment (SCOE) for the test and measurement of communication satellite Payload at subsystem and system level. The main emphasis of this paper is to demonstrate the principle test equipment, instruments and the payload test matrix for an automatic test control. Electrical Ground Support Equipment (EGSE)/ Special Check out Equipment (SCOE) requirements, functions and architecture for C-band and Ku-band payloads are presented in details along with their interface with satellite during different phases of satellite testing. It provides test setup, in a single rack cabinet that can easily be moved from payload assembly and integration environment to thermal vacuum chamber all the way to launch site (for pre-launch test and verification).

  18. SU-E-J-88: The Study of Setup Error Measured by CBCT in Postoperative Radiotherapy for Cervical Carcinoma

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Runxiao, L; Aikun, W; Xiaomei, F

    2015-06-15

    Purpose: To compare two registration methods in the CBCT guided radiotherapy for cervical carcinoma, analyze the setup errors and registration methods, determine the margin required for clinical target volume(CTV) extending to planning target volume(PTV). Methods: Twenty patients with cervical carcinoma were enrolled. All patients were underwent CT simulation in the supine position. Transfering the CT images to the treatment planning system and defining the CTV, PTV and the organs at risk (OAR), then transmit them to the XVI workshop. CBCT scans were performed before radiotherapy and registered to planning CT images according to bone and gray value registration methods. Comparedmore » two methods and obtain left-right(X), superior-inferior(Y), anterior-posterior (Z) setup errors, the margin required for CTV to PTV were calculated. Results: Setup errors were unavoidable in postoperative cervical carcinoma irradiation. The setup errors measured by method of bone (systemic ± random) on X(1eft.right),Y(superior.inferior),Z(anterior.posterior) directions were(0.24±3.62),(0.77±5.05) and (0.13±3.89)mm, respectively, the setup errors measured by method of grey (systemic ± random) on X(1eft-right), Y(superior-inferior), Z(anterior-posterior) directions were(0.31±3.93), (0.85±5.16) and (0.21±4.12)mm, respectively.The spatial distributions of setup error was maximum in Y direction. The margins were 4 mm in X axis, 6 mm in Y axis, 4 mm in Z axis respectively.These two registration methods were similar and highly recommended. Conclusion: Both bone and grey registration methods could offer an accurate setup error. The influence of setup errors of a PTV margin would be suggested by 4mm, 4mm and 6mm on X, Y and Z directions for postoperative radiotherapy for cervical carcinoma.« less

  19. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rong, Y; Walston, S

    Purpose: To evaluate the use of 3D optical surface imaging as a new surrogate for respiratory motion gated deep-inspiration breath-hold (DIBH) technique for left breast cancer patients. Methods: Patients with left-sided breast cancer after lumpectomy or mastectomy were selected as candidates for DIBH technique for their external beam radiation therapy. Treatment plans were created on both free breathing (FB) and DIBH CTs to determine whether DIBH was beneficial in reducing heart doses. The Real-time Position Management (RPM) system was used to acquire patient's breathing trace during DIBH CT acquisition and treatment delivery. The reference 3D surface models from FB andmore » DIBH CTs were generated and transferred to the “AlignRT” system for patient positioning and real-time treatment monitoring. MV Cine images were acquired for each beam as quality assurance for intra-fractional position verification. The chest wall excursions measured on these images were used to define the actual target position during treatment, and to investigate the accuracy and reproducibility of RPM and AlignRT. Results: Reduction in heart dose can be achieved for left-sided breast patients using DIBH. Results showed that RPM has poor correlation with target position, as determined by the MV Cine imaging. This indicates that RPM may not be an adequate surrogate in defining the breath-hold level when used alone. Alternatively, the AlignRT surface imaging demonstrated a better correlation with the actual CW excursion during DIBH. Both the vertical and magnitude real-time deltas (RTDs) reported by AlignRT can be used as the gating parameter, with a recommend threshold of ±3 mm and 5 mm, respectively. Conclusion: 3D optical surface imaging serves as a superior target surrogate for the left breast treatment when compared to RPM. Working together with the realtime MV Cine imaging, they ensure accurate patient setup and dose delivery, while minimizing the imaging dose to patients.« less

  20. Cone-Beam CT Assessment of Interfraction and Intrafraction Setup Error of Two Head-and-Neck Cancer Thermoplastic Masks

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Velec, Michael; Waldron, John N.; O'Sullivan, Brian

    2010-03-01

    Purpose: To prospectively compare setup error in standard thermoplastic masks and skin-sparing masks (SSMs) modified with low neck cutouts for head-and-neck intensity-modulated radiation therapy (IMRT) patients. Methods and Materials: Twenty head-and-neck IMRT patients were randomized to be treated in a standard mask (SM) or SSM. Cone-beam computed tomography (CBCT) scans, acquired daily after both initial setup and any repositioning, were used for initial and residual interfraction evaluation, respectively. Weekly, post-IMRT CBCT scans were acquired for intrafraction setup evaluation. The population random (sigma) and systematic (SIGMA) errors were compared for SMs and SSMs. Skin toxicity was recorded weekly by use ofmore » Radiation Therapy Oncology Group criteria. Results: We evaluated 762 CBCT scans in 11 patients randomized to the SM and 9 to the SSM. Initial interfraction sigma was 1.6 mm or less or 1.1 deg. or less for SM and 2.0 mm or less and 0.8 deg. for SSM. Initial interfraction SIGMA was 1.0 mm or less or 1.4 deg. or less for SM and 1.1 mm or less or 0.9 deg. or less for SSM. These errors were reduced before IMRT with CBCT image guidance with no significant differences in residual interfraction or intrafraction uncertainties between SMs and SSMs. Intrafraction sigma and SIGMA were less than 1 mm and less than 1 deg. for both masks. Less severe skin reactions were observed in the cutout regions of the SSM compared with non-cutout regions. Conclusions: Interfraction and intrafraction setup error is not significantly different for SSMs and conventional masks in head-and-neck radiation therapy. Mask cutouts should be considered for these patients in an effort to reduce skin toxicity.« less

  1. Assuring high quality treatment delivery in clinical trials - Results from the Trans-Tasman Radiation Oncology Group (TROG) study 03.04 "RADAR" set-up accuracy study.

    PubMed

    Haworth, Annette; Kearvell, Rachel; Greer, Peter B; Hooton, Ben; Denham, James W; Lamb, David; Duchesne, Gillian; Murray, Judy; Joseph, David

    2009-03-01

    A multi-centre clinical trial for prostate cancer patients provided an opportunity to introduce conformal radiotherapy with dose escalation. To verify adequate treatment accuracy prior to patient recruitment, centres submitted details of a set-up accuracy study (SUAS). We report the results of the SUAS, the variation in clinical practice and the strategies used to help centres improve treatment accuracy. The SUAS required each of the 24 participating centres to collect data on at least 10 pelvic patients imaged on a minimum of 20 occasions. Software was provided for data collection and analysis. Support to centres was provided through educational lectures, the trial quality assurance team and an information booklet. Only two centres had recently carried out a SUAS prior to the trial opening. Systematic errors were generally smaller than those previously reported in the literature. The questionnaire identified many differences in patient set-up protocols. As a result of participating in this QA activity more than 65% of centres improved their treatment delivery accuracy. Conducting a pre-trial SUAS has led to improvement in treatment delivery accuracy in many centres. Treatment techniques and set-up accuracy varied greatly, demonstrating a need to ensure an on-going awareness for such studies in future trials and with the introduction of dose escalation or new technologies.

  2. SU-F-T-536: Contra-Lateral Breast Study for Prone Versus Supine Patients

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Marrero, M; Joseph, K; Klein, E

    Purpose: There are several advantages to utilizing the prone technique for intact breast cancer patients. However, as the topography changes, accompanied by the influence of a supporting breast board and patient treatment couch, the question that arises is to whether there is a concern for contralateral breast dose for intact breast cancer patients being treated with this technique. Methods: An anthropomorphic phantom with breast mounds to duplicate intact breast cancer treatment was planned in prone and supine position. Two tangential beams were executed in the similar manner for as the radiotherapy planning system. For the prone setup, a breast densemore » foam board was used to support the phantom. A grid of 24 OSL nanodots was placed at 6cm, 4cm, and 2cm apart from the medial border for both prone and supine setups. The phantom was set up using megavoltage imaging and treated as per plan. Additional, a similar study was performed on a patient treated in prone position. Results: Overall, the contralateral breast dose was generally higher for prone setups at all locations especially when close to the medial border. The average mean dose was found to be 1.8%, 2.5% of the prescribed dose for supine respectively prone position. The average of the standard deviation is 1.04%, 1.38% for supine respectively prone position. As for patient treated in prone position average mean dose was found to be 1.165% of the prescribed dose and average of the standard deviation is 9.456%. Conclusion: There is minimal influence of scatter from the breast board. It appears that the volatility of the setup could lead to higher doses than expected from the planning system to the contralateral breast when the patient is in the prone position.« less

  3. BrainCycles: Experimental Setup for the Combined Measurement of Cortical and Subcortical Activity in Parkinson's Disease Patients during Cycling.

    PubMed

    Gratkowski, Maciej; Storzer, Lena; Butz, Markus; Schnitzler, Alfons; Saupe, Dietmar; Dalal, Sarang S

    2016-01-01

    Recently, it has been demonstrated that bicycling ability remains surprisingly preserved in Parkinson's disease (PD) patients who suffer from freezing of gait. Cycling has been also proposed as a therapeutic means of treating PD symptoms, with some preliminary success. The neural mechanisms behind these phenomena are however not yet understood. One of the reasons is that the investigations of neuronal activity during pedaling have been up to now limited to PET and fMRI studies, which restrict the temporal resolution of analysis, and to scalp EEG focused on cortical activation. However, deeper brain structures like the basal ganglia are also associated with control of voluntary motor movements like cycling and are affected by PD. Deep brain stimulation (DBS) electrodes implanted for therapy in PD patients provide rare and unique access to directly record basal ganglia activity with a very high temporal resolution. In this paper we present an experimental setup allowing combined investigation of basal ganglia local field potentials (LFPs) and scalp EEG underlying bicycling in PD patients. The main part of the setup is a bike simulator consisting of a classic Dutch-style bicycle frame mounted on a commercially available ergometer. The pedal resistance is controllable in real-time by custom software and the pedal position is continuously tracked by custom Arduino-based electronics using optical and magnetic sensors. A portable bioamplifier records the pedal position signal, the angle of the knee, and the foot pressure together with EEG, EMG, and basal ganglia LFPs. A handlebar-mounted display provides additional information for patients riding the bike simulator, including the current and target pedaling rate. In order to demonstrate the utility of the setup, example data from pilot recordings are shown. The presented experimental setup provides means to directly record basal ganglia activity not only during cycling but also during other movement tasks in patients who have undergone DBS treatment. Thus, it can facilitate studies comparing bicycling and walking, to elucidate why PD patients often retain the ability to bicycle despite severe freezing of gait. Moreover it can help clarifying the mechanism through which cycling may have therapeutic benefits.

  4. BrainCycles: Experimental Setup for the Combined Measurement of Cortical and Subcortical Activity in Parkinson's Disease Patients during Cycling

    PubMed Central

    Gratkowski, Maciej; Storzer, Lena; Butz, Markus; Schnitzler, Alfons; Saupe, Dietmar; Dalal, Sarang S.

    2017-01-01

    Recently, it has been demonstrated that bicycling ability remains surprisingly preserved in Parkinson's disease (PD) patients who suffer from freezing of gait. Cycling has been also proposed as a therapeutic means of treating PD symptoms, with some preliminary success. The neural mechanisms behind these phenomena are however not yet understood. One of the reasons is that the investigations of neuronal activity during pedaling have been up to now limited to PET and fMRI studies, which restrict the temporal resolution of analysis, and to scalp EEG focused on cortical activation. However, deeper brain structures like the basal ganglia are also associated with control of voluntary motor movements like cycling and are affected by PD. Deep brain stimulation (DBS) electrodes implanted for therapy in PD patients provide rare and unique access to directly record basal ganglia activity with a very high temporal resolution. In this paper we present an experimental setup allowing combined investigation of basal ganglia local field potentials (LFPs) and scalp EEG underlying bicycling in PD patients. The main part of the setup is a bike simulator consisting of a classic Dutch-style bicycle frame mounted on a commercially available ergometer. The pedal resistance is controllable in real-time by custom software and the pedal position is continuously tracked by custom Arduino-based electronics using optical and magnetic sensors. A portable bioamplifier records the pedal position signal, the angle of the knee, and the foot pressure together with EEG, EMG, and basal ganglia LFPs. A handlebar-mounted display provides additional information for patients riding the bike simulator, including the current and target pedaling rate. In order to demonstrate the utility of the setup, example data from pilot recordings are shown. The presented experimental setup provides means to directly record basal ganglia activity not only during cycling but also during other movement tasks in patients who have undergone DBS treatment. Thus, it can facilitate studies comparing bicycling and walking, to elucidate why PD patients often retain the ability to bicycle despite severe freezing of gait. Moreover it can help clarifying the mechanism through which cycling may have therapeutic benefits. PMID:28119591

  5. Effect of Body Mass Index on Magnitude of Setup Errors in Patients Treated With Adjuvant Radiotherapy for Endometrial Cancer With Daily Image Guidance

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lin, Lilie L., E-mail: lin@uphs.upenn.edu; Hertan, Lauren; Rengan, Ramesh

    2012-06-01

    Purpose: To determine the impact of body mass index (BMI) on daily setup variations and frequency of imaging necessary for patients with endometrial cancer treated with adjuvant intensity-modulated radiotherapy (IMRT) with daily image guidance. Methods and Materials: The daily shifts from a total of 782 orthogonal kilovoltage images from 30 patients who received pelvic IMRT between July 2008 and August 2010 were analyzed. The BMI, mean daily shifts, and random and systematic errors in each translational and rotational direction were calculated for each patient. Margin recipes were generated based on BMI. Linear regression and spearman rank correlation analysis were performed.more » To simulate a less-than-daily IGRT protocol, the average shift of the first five fractions was applied to subsequent setups without IGRT for assessing the impact on setup error and margin requirements. Results: Median BMI was 32.9 (range, 23-62). Of the 30 patients, 16.7% (n = 5) were normal weight (BMI <25); 23.3% (n = 7) were overweight (BMI {>=}25 to <30); 26.7% (n = 8) were mildly obese (BMI {>=}30 to <35); and 33.3% (n = 10) were moderately to severely obese (BMI {>=} 35). On linear regression, mean absolute vertical, longitudinal, and lateral shifts positively correlated with BMI (p = 0.0127, p = 0.0037, and p < 0.0001, respectively). Systematic errors in the longitudinal and vertical direction were found to be positively correlated with BMI category (p < 0.0001 for both). IGRT for the first five fractions, followed by correction of the mean error for all subsequent fractions, led to a substantial reduction in setup error and resultant margin requirement overall compared with no IGRT. Conclusions: Daily shifts, systematic errors, and margin requirements were greatest in obese patients. For women who are normal or overweight, a planning target margin margin of 7 to 10 mm may be sufficient without IGRT, but for patients who are moderately or severely obese, this is insufficient.« less

  6. Real-Time Verification of a High-Dose-Rate Iridium 192 Source Position Using a Modified C-Arm Fluoroscope

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Nose, Takayuki, E-mail: nose-takayuki@nms.ac.jp; Chatani, Masashi; Otani, Yuki

    Purpose: High-dose-rate (HDR) brachytherapy misdeliveries can occur at any institution, and they can cause disastrous results. Even a patient's death has been reported. Misdeliveries could be avoided with real-time verification methods. In 1996, we developed a modified C-arm fluoroscopic verification of an HDR Iridium 192 source position prevent these misdeliveries. This method provided excellent image quality sufficient to detect errors, and it has been in clinical use at our institutions for 20 years. The purpose of the current study is to introduce the mechanisms and validity of our straightforward C-arm fluoroscopic verification method. Methods and Materials: Conventional X-ray fluoroscopic images aremore » degraded by spurious signals and quantum noise from Iridium 192 photons, which make source verification impractical. To improve image quality, we quadrupled the C-arm fluoroscopic X-ray dose per pulse. The pulse rate was reduced by a factor of 4 to keep the average exposure compliant with Japanese medical regulations. The images were then displayed with quarter-frame rates. Results: Sufficient quality was obtained to enable observation of the source position relative to both the applicators and the anatomy. With this method, 2 errors were detected among 2031 treatment sessions for 370 patients within a 6-year period. Conclusions: With the use of a modified C-arm fluoroscopic verification method, treatment errors that were otherwise overlooked were detected in real time. This method should be given consideration for widespread use.« less

  7. Real-Time Verification of a High-Dose-Rate Iridium 192 Source Position Using a Modified C-Arm Fluoroscope.

    PubMed

    Nose, Takayuki; Chatani, Masashi; Otani, Yuki; Teshima, Teruki; Kumita, Shinichirou

    2017-03-15

    High-dose-rate (HDR) brachytherapy misdeliveries can occur at any institution, and they can cause disastrous results. Even a patient's death has been reported. Misdeliveries could be avoided with real-time verification methods. In 1996, we developed a modified C-arm fluoroscopic verification of an HDR Iridium 192 source position prevent these misdeliveries. This method provided excellent image quality sufficient to detect errors, and it has been in clinical use at our institutions for 20 years. The purpose of the current study is to introduce the mechanisms and validity of our straightforward C-arm fluoroscopic verification method. Conventional X-ray fluoroscopic images are degraded by spurious signals and quantum noise from Iridium 192 photons, which make source verification impractical. To improve image quality, we quadrupled the C-arm fluoroscopic X-ray dose per pulse. The pulse rate was reduced by a factor of 4 to keep the average exposure compliant with Japanese medical regulations. The images were then displayed with quarter-frame rates. Sufficient quality was obtained to enable observation of the source position relative to both the applicators and the anatomy. With this method, 2 errors were detected among 2031 treatment sessions for 370 patients within a 6-year period. With the use of a modified C-arm fluoroscopic verification method, treatment errors that were otherwise overlooked were detected in real time. This method should be given consideration for widespread use. Copyright © 2016 Elsevier Inc. All rights reserved.

  8. Improving the modelling of irradiation-induced brain activation for in vivo PET verification of proton therapy.

    PubMed

    Bauer, Julia; Chen, Wenjing; Nischwitz, Sebastian; Liebl, Jakob; Rieken, Stefan; Welzel, Thomas; Debus, Juergen; Parodi, Katia

    2018-04-24

    A reliable Monte Carlo prediction of proton-induced brain tissue activation used for comparison to particle therapy positron-emission-tomography (PT-PET) measurements is crucial for in vivo treatment verification. Major limitations of current approaches to overcome include the CT-based patient model and the description of activity washout due to tissue perfusion. Two approaches were studied to improve the activity prediction for brain irradiation: (i) a refined patient model using tissue classification based on MR information and (ii) a PT-PET data-driven refinement of washout model parameters. Improvements of the activity predictions compared to post-treatment PT-PET measurements were assessed in terms of activity profile similarity for six patients treated with a single or two almost parallel fields delivered by active proton beam scanning. The refined patient model yields a generally higher similarity for most of the patients, except in highly pathological areas leading to tissue misclassification. Using washout model parameters deduced from clinical patient data could considerably improve the activity profile similarity for all patients. Current methods used to predict proton-induced brain tissue activation can be improved with MR-based tissue classification and data-driven washout parameters, thus providing a more reliable basis for PT-PET verification. Copyright © 2018 Elsevier B.V. All rights reserved.

  9. Proton radiography for inline treatment planning and positioning verification of small animals.

    PubMed

    Müller, Johannes; Neubert, Christian; von Neubeck, Cläre; Baumann, Michael; Krause, Mechthild; Enghardt, Wolfgang; Bütof, Rebecca; Dietrich, Antje; Lühr, Armin

    2017-11-01

    As proton therapy becomes increasingly well established, there is a need for high-quality clinically relevant in vivo data to gain better insight into the radiobiological effects of proton irradiation on both healthy and tumor tissue. This requires the development of easily applicable setups that allow for efficient, fractionated, image-guided proton irradiation of small animals, the most widely used pre-clinical model. Here, a method is proposed to perform dual-energy proton radiography for inline positioning verification and treatment planning. Dual-energy proton radiography exploits the differential enhancement of object features in two successively measured two-dimensional (2D) dose distributions at two different proton energies. The two raw images show structures that are dominated by energy absorption (absorption mode) or scattering (scattering mode) of protons in the object, respectively. Data post-processing allowed for the separation of both signal contributions in the respective images. The images were evaluated regarding recognizable object details and feasibility of rigid registration to acquired planar X-ray scans. Robust, automated rigid registration of proton radiography and planar X-ray images in scattering mode could be reliably achieved with the animal bedding unit used as registration landmark. Distinguishable external and internal features of the imaged mouse included the outer body contour, the skull with substructures, the lung, abdominal structures and the hind legs. Image analysis based on the combined information of both imaging modes allowed image enhancement and calculation of 2D water-equivalent path length (WEPL) maps of the object along the beam direction. Fractionated irradiation of exposed target volumes (e.g., subcutaneous tumor model or brain) can be realized with the suggested method being used for daily positioning and range determination. Robust registration of X-ray and proton radiography images allows for the irradiation of tumor entities that require conventional computed tomography (CT)-based planning, such as orthotopic lung or brain tumors, similar to conventional patient treatment.

  10. SU-E-J-53: A Phantom Design to Assist Patient Position Verification System in Daily Image-Guided RT and Comprehensive QA Measurements

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Syh, J; Wu, H

    2015-06-15

    Purpose This study is to implement a homemade novel device with surface locking couch index to check daily radiograph (DR) function of adaPTInsight™, stereoscopic image guided system (SIGS), for proton therapy. The comprehensive daily QA checks of proton pencil beam output, field size, flatness and symmetry of spots and energy layers will be followed by using MatriXX dosimetry device. Methods The iBa MatriXX device was used to perform daily dosimetry which is also used to perform SIGS checks. A set of markers were attached to surface of MatriXX device in alignment of DRR of reconstructed CT images and daily DR.more » The novel device allows MatriXX to be fit into the cradle which was locked by couch index bars on couch surface. This will keep the MatriXX at same XY plane daily with exact coordinates. Couch height Z will be adjusted according to imaging to check isocenter-laser coincidence accuracy. Results adaPTInsight™ provides robotic couch to move in 6-degree coordinate system to align the dosimetry device to be within 1.0 mm / 1.0°. The daily constancy was tightened to be ± 0.5 mm / 0.3° compared to 1.0 mm / 1.0° before. For gantry at 0° and couch all 0° angles (@ Rt ARM 0 setting), offsets measured of the couch systems were ≤ 0.5° in roll, yaw and pitch dimensions. Conclusion Simplicity of novel device made daily image guided QA consistent with accuracy. The offset of the MatriXX isocenter-laser coincident was reproducible. Such easy task not only speeds up the setup, but it increases confidence level in detailed daily comprehensive measurements. The total SIGS alignment time has been shortened with less setup error. This device will enhance our experiences for the future QA when cone beam CT imaging modality becomes available at proton therapy center.« less

  11. SU-F-T-92: Clinical Benefit for Breast and Chest Wall Setup in Using a Breast Board

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Li, S; Miyamoto, C; Serratore, D

    Purpose: To validate benefit of using a breast board (BB) by analyzing the geometry and dosimetry changes of the regions of interest (ROIs) between CT scans with and without BB. Methods: Seven patients, two chest walls (CW) and five breasts, use BB at CT simulation and no BB at diagnostic CT were included. By using deformable image registration software (Velocity AI), diagnostic CT and planning CT were rigidly co-registered according to the thoracic cage at the target. The heart and the target were then deformedly matched and the contours of the planned ROIs were transferred to the diagnostic CT. Whichmore » were brought back to the planning CT data set though the initial rigid co-registration in order to keep the deformed ROIs redefined in the diagnostic CT. Anatomic shifts and volume changes of a ROI beyond the rigid translation were recorded and dosimetry changes to ROIs were compared with recalculated DVHs. Results: Patient setup without the BB had small but systematic heart shifts superiorly by ∼5 mm. Torso rotations in two cases moved the heart in opposite directions by ∼10 mm. The breast target volume, shape, and locations were significantly changed with arm extension over the head but not in cases with the arm extended laterally. Breast setup without BB could increase the mean dose to the heart and the maximal dose to the anterior ventricle wall by 1.1 and 6.7 Gy, respectively. Conclusion: A method for evaluation of breast setup technique is introduced and applied for patients. Results of systematic heart displacement without using the BB and the potential increase of heart doses encourage us to further investigate the current trend of not using a BB for easy setup and CT scans. Using a BB would likely increase patient sag during prolonged IMRT and real-time patient position monitoring is clinically desired.« less

  12. Verification of performance specifications for a US Food and Drug Administration-approved molecular microbiology test: Clostridium difficile cytotoxin B using the Becton, Dickinson and Company GeneOhm Cdiff assay.

    PubMed

    Schlaberg, Robert; Mitchell, Michael J; Taggart, Edward W; She, Rosemary C

    2012-01-01

    US Food and Drug Administration (FDA)-approved diagnostic tests based on molecular genetic technologies are becoming available for an increasing number of microbial pathogens. Advances in technology and lower costs have moved molecular diagnostic tests formerly performed for research purposes only into much wider use in clinical microbiology laboratories. To provide an example of laboratory studies performed to verify the performance of an FDA-approved assay for the detection of Clostridium difficile cytotoxin B compared with the manufacturer's performance standards. We describe the process and protocols used by a laboratory for verification of an FDA-approved assay, assess data from the verification studies, and implement the assay after verification. Performance data from the verification studies conducted by the laboratory were consistent with the manufacturer's performance standards and the assay was implemented into the laboratory's test menu. Verification studies are required for FDA-approved diagnostic assays prior to use in patient care. Laboratories should develop a standardized approach to verification studies that can be adapted and applied to different types of assays. We describe the verification of an FDA-approved real-time polymerase chain reaction assay for the detection of a toxin gene in a bacterial pathogen.

  13. Improving Patient Safety With Error Identification in Chemotherapy Orders by Verification Nurses.

    PubMed

    Baldwin, Abigail; Rodriguez, Elizabeth S

    2016-02-01

    The prevalence of medication errors associated with chemotherapy administration is not precisely known. Little evidence exists concerning the extent or nature of errors; however, some evidence demonstrates that errors are related to prescribing. This article demonstrates how the review of chemotherapy orders by a designated nurse known as a verification nurse (VN) at a National Cancer Institute-designated comprehensive cancer center helps to identify prescribing errors that may prevent chemotherapy administration mistakes and improve patient safety in outpatient infusion units. This article will describe the role of the VN and details of the verification process. To identify benefits of the VN role, a retrospective review and analysis of chemotherapy near-miss events from 2009-2014 was performed. A total of 4,282 events related to chemotherapy were entered into the Reporting to Improve Safety and Quality system. A majority of the events were categorized as near-miss events, or those that, because of chance, did not result in patient injury, and were identified at the point of prescribing.

  14. Length matters: Improved high field EEG-fMRI recordings using shorter EEG cables.

    PubMed

    Assecondi, Sara; Lavallee, Christina; Ferrari, Paolo; Jovicich, Jorge

    2016-08-30

    The use of concurrent EEG-fMRI recordings has increased in recent years, allowing new avenues of medical and cognitive neuroscience research; however, currently used setups present problems with data quality and reproducibility. We propose a compact experimental setup for concurrent EEG-fMRI at 4T and compare it to a more standard reference setup. The compact setup uses short EEG cables connecting to the amplifiers, which are placed right at the back of the head RF coil on a form-fitting extension force-locked to the patient MR bed. We compare the two setups in terms of sensitivity to MR-room environmental noise, interferences between measuring devices (EEG or fMRI), and sensitivity to functional responses in a visual stimulation paradigm. The compact setup reduces the system sensitivity to both external noise and MR-induced artefacts by at least 60%, with negligible EEG noise induced from the mechanical vibrations of the cryogenic cooling compression pump. The compact setup improved EEG data quality and the overall performance of MR-artifact correction techniques. Both setups were similar in terms of the fMRI data, with higher reproducibility for cable placement within the scanner in the compact setup. This improved compact setup may be relevant to MR laboratories interested in reducing the sensitivity of their EEG-fMRI experimental setup to external noise sources, setting up an EEG-fMRI workplace for the first time, or for creating a more reproducible configuration of equipment and cables. Implications for safety and ergonomics are discussed. Copyright © 2016 Elsevier B.V. All rights reserved.

  15. Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

    NASA Astrophysics Data System (ADS)

    Tyagi, N.; Curran, B. H.; Roberson, P. L.; Moran, J. M.; Acosta, E.; Fraass, B. A.

    2008-02-01

    IMRT often requires delivering small fields which may suffer from electronic disequilibrium effects. The presence of heterogeneities, particularly low-density tissues in patients, complicates such situations. In this study, we report on verification of the DPM MC code for IMRT treatment planning in heterogeneous media, using a previously developed model of the Varian 120-leaf MLC. The purpose of this study is twofold: (a) design a comprehensive list of experiments in heterogeneous media for verification of any dose calculation algorithm and (b) verify our MLC model in these heterogeneous type geometries that mimic an actual patient geometry for IMRT treatment. The measurements have been done using an IMRT head and neck phantom (CIRS phantom) and slab phantom geometries. Verification of the MLC model has been carried out using point doses measured with an A14 slim line (SL) ion chamber inside a tissue-equivalent and a bone-equivalent material using the CIRS phantom. Planar doses using lung and bone equivalent slabs have been measured and compared using EDR films (Kodak, Rochester, NY).

  16. Experimental Evaluation of Verification and Validation Tools on Martian Rover Software

    NASA Technical Reports Server (NTRS)

    Brat, Guillaume; Giannakopoulou, Dimitra; Goldberg, Allen; Havelund, Klaus; Lowry, Mike; Pasareani, Corina; Venet, Arnaud; Visser, Willem; Washington, Rich

    2003-01-01

    We report on a study to determine the maturity of different verification and validation technologies (V&V) on a representative example of NASA flight software. The study consisted of a controlled experiment where three technologies (static analysis, runtime analysis and model checking) were compared to traditional testing with respect to their ability to find seeded errors in a prototype Mars Rover. What makes this study unique is that it is the first (to the best of our knowledge) to do a controlled experiment to compare formal methods based tools to testing on a realistic industrial-size example where the emphasis was on collecting as much data on the performance of the tools and the participants as possible. The paper includes a description of the Rover code that was analyzed, the tools used as well as a detailed description of the experimental setup and the results. Due to the complexity of setting up the experiment, our results can not be generalized, but we believe it can still serve as a valuable point of reference for future studies of this kind. It did confirm the belief we had that advanced tools can outperform testing when trying to locate concurrency errors. Furthermore the results of the experiment inspired a novel framework for testing the next generation of the Rover.

  17. TU-H-CAMPUS-TeP1-02: Seated Treatment: Setup Uncertainty Comparable to Supine

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    McCarroll, R; UT Health Science Center, Graduate School of Biomedical Sciences, Houston, TX; Beadle, B

    Purpose: For some head and neck patients, positioning in the supine position is not well tolerated. For these patients, treatment in a seated position would be preferred. We have evaluated inter- and intra- fraction uncertainty of patient set-up in a novel treatment chair which is compatible with modern linac designs. Methods: Five head-and-neck cancer patients were positioned in the chair, fitted with immobilization devices, and imaged with orthogonal X-rays. The couch (with chair attached) was rotated to simulate delivery (without actual treatment), another set of images were acquired, providing a measure of intra-fraction displacement. The patient then got off ofmore » and back onto the chair and the process was repeated, thus providing a measure of inter-fraction set-up uncertainty. Six sub-regions in the head-and-neck were rigidly registered to evaluate local intra- and interfraction displacement. Image guidance was simulated by first registering one sub-region; the residual displacement of other sub-regions was then measured. Additionally, a patient questionnaire was administered to evaluate tolerance of the seated position. Results: The chair design is such that all advantages of couch motions may be utilized. Average inter- and intrafraction displacements of all sub-regions in the seated position were less than 2 and 3 mm, respectively. When image guidance was simulated, interfraction displacements were reduced by an average of 4 mm, providing comparable setup to the supine position. The enrolled patients, who had no indication for a seated treatment position, reported no preference for the seated or the supine position. Conclusion: The novel chair design provides acceptable inter- and intra-fraction displacement, with reproducibility similar to that observed for patients in the supine position. Such a chair will be utilized for patients who cannot tolerate the supine position and use with CBCT images for planning, in a fixed-beam linac system, and for other treatment sites is under investigation. Funding: Varian Medical Systems.« less

  18. Experimental and Numerical Studies on Wave Breaking Characteristics over a Fringing Reef under Monochromatic Wave Conditions

    PubMed Central

    2014-01-01

    Fringing reefs play an important role in protecting the coastal area by inducing wave breaking and wave energy dissipation. However, modeling of wave transformation and energy dissipation on this topography is still difficult due to the unique structure. In the present study, two-dimensional laboratory experiments were conducted to investigate the cross-shore variations of wave transformation, setup, and breaking phenomena over an idealized fringing reef with the 1/40 reef slope and to verify the Boussinesq model under monochromatic wave conditions. One-layer and two-layer model configurations of the Boussinesq model were used to figure out the model capability. Both models predicted well (r 2 > 0.8) the cross-shore variation of the wave heights, crests, troughs, and setups when the nonlinearity is not too high (A 0/h 0 < 0.07 in this study). However, as the wave nonlinearity and steepness increase, the one-layer model showed problems in prediction and stability due to the error on the vertical profile of fluid velocity. The results in this study revealed that one-layer model is not suitable in the highly nonlinear wave condition over a fringing reef bathymetry. This data set can contribute to the numerical model verification. PMID:25276853

  19. Explosive component acceptance tester using laser interferometer technology

    NASA Technical Reports Server (NTRS)

    Wickstrom, Richard D.; Tarbell, William W.

    1993-01-01

    Acceptance testing of explosive components requires a reliable and simple to use testing method that can discern less than optimal performance. For hot-wire detonators, traditional techniques use dent blocks or photographic diagnostic methods. More complicated approaches are avoided because of their inherent problems with setup and maintenance. A recently developed tester is based on using a laser interferometer to measure the velocity of flying plates accelerated by explosively actuated detonators. Unlike ordinary interferometers that monitor displacement of the test article, this device measures velocity directly and is commonly used with non-spectral surfaces. Most often referred to as the VISAR technique (Velocity Interferometer System for Any Reflecting Surface), it has become the most widely-accepted choice for accurate measurement of velocity in the range greater than 1 mm/micro-s. Traditional VISAR devices require extensive setup and adjustment and therefore are unacceptable in a production-testing environment. This paper describes a new VISAR approach which requires virtually no adjustments, yet provides data with accuracy comparable to the more complicated systems. The device, termed the Fixed-Cavity VISAR, is currently being developed to serve as a product verification tool for hot-wire detonators and slappers. An extensive data acquisition and analysis computer code was also created to automate the manipulation of raw data into final results.

  20. Experimental and numerical studies on wave breaking characteristics over a fringing reef under monochromatic wave conditions.

    PubMed

    Lee, Jong-In; Shin, Sungwon; Kim, Young-Taek

    2014-01-01

    Fringing reefs play an important role in protecting the coastal area by inducing wave breaking and wave energy dissipation. However, modeling of wave transformation and energy dissipation on this topography is still difficult due to the unique structure. In the present study, two-dimensional laboratory experiments were conducted to investigate the cross-shore variations of wave transformation, setup, and breaking phenomena over an idealized fringing reef with the 1/40 reef slope and to verify the Boussinesq model under monochromatic wave conditions. One-layer and two-layer model configurations of the Boussinesq model were used to figure out the model capability. Both models predicted well (r (2) > 0.8) the cross-shore variation of the wave heights, crests, troughs, and setups when the nonlinearity is not too high (A 0/h 0 < 0.07 in this study). However, as the wave nonlinearity and steepness increase, the one-layer model showed problems in prediction and stability due to the error on the vertical profile of fluid velocity. The results in this study revealed that one-layer model is not suitable in the highly nonlinear wave condition over a fringing reef bathymetry. This data set can contribute to the numerical model verification.

  1. Reduction of prostate intrafraction motion using gas-release rectal balloons

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Su Zhong; Zhao Tianyu; Li Zuofeng

    2012-10-15

    Purpose: To analyze prostate intrafraction motion using both non-gas-release (NGR) and gas-release (GR) rectal balloons and to evaluate the ability of GR rectal balloons to reduce prostate intrafraction motion. Methods: Twenty-nine patients with NGR rectal balloons and 29 patients with GR balloons were randomly selected from prostate patients treated with proton therapy at University of Florida Proton Therapy Institute (Jacksonville, FL). Their pretreatment and post-treatment orthogonal radiographs were analyzed, and both pretreatment setup residual error and intrafraction-motion data were obtained. Population histograms of intrafraction motion were plotted for both types of balloons. Population planning target-volume (PTV) margins were calculated withmore » the van Herk formula of 2.5{Sigma}+ 0.7{sigma} to account for setup residual errors and intrafraction motion errors. Results: Pretreatment and post-treatment radiographs indicated that the use of gas-release rectal balloons reduced prostate intrafraction motion along superior-inferior (SI) and anterior-posterior (AP) directions. Similar patient setup residual errors were exhibited for both types of balloons. Gas-release rectal balloons resulted in PTV margin reductions from 3.9 to 2.8 mm in the SI direction, 3.1 to 1.8 mm in the AP direction, and an increase from 1.9 to 2.1 mm in the left-right direction. Conclusions: Prostate intrafraction motion is an important uncertainty source in radiotherapy after image-guided patient setup with online corrections. Compared to non-gas-release rectal balloons, gas-release balloons can reduce prostate intrafraction motion in the SI and AP directions caused by gas buildup.« less

  2. A Voluntary Breath-Hold Treatment Technique for the Left Breast With Unfavorable Cardiac Anatomy Using Surface Imaging

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Gierga, David P., E-mail: dgierga@partners.org; Harvard Medical School, Boston, Massachusetts; Turcotte, Julie C.

    2012-12-01

    Purpose: Breath-hold (BH) treatments can be used to reduce cardiac dose for patients with left-sided breast cancer and unfavorable cardiac anatomy. A surface imaging technique was developed for accurate patient setup and reproducible real-time BH positioning. Methods and Materials: Three-dimensional surface images were obtained for 20 patients. Surface imaging was used to correct the daily setup for each patient. Initial setup data were recorded for 443 fractions and were analyzed to assess random and systematic errors. Real time monitoring was used to verify surface placement during BH. The radiation beam was not turned on if the BH position difference wasmore » greater than 5 mm. Real-time surface data were analyzed for 2398 BHs and 363 treatment fractions. The mean and maximum differences were calculated. The percentage of BHs greater than tolerance was calculated. Results: The mean shifts for initial patient setup were 2.0 mm, 1.2 mm, and 0.3 mm in the vertical, longitudinal, and lateral directions, respectively. The mean 3-dimensional vector shift was 7.8 mm. Random and systematic errors were less than 4 mm. Real-time surface monitoring data indicated that 22% of the BHs were outside the 5-mm tolerance (range, 7%-41%), and there was a correlation with breast volume. The mean difference between the treated and reference BH positions was 2 mm in each direction. For out-of-tolerance BHs, the average difference in the BH position was 6.3 mm, and the average maximum difference was 8.8 mm. Conclusions: Daily real-time surface imaging ensures accurate and reproducible positioning for BH treatment of left-sided breast cancer patients with unfavorable cardiac anatomy.« less

  3. Investigation of the optimum location of external markers for patient setup accuracy enhancement at external beam radiotherapy

    PubMed Central

    Torshabi, Ahmad Esmaili; Nankali, Saber

    2016-01-01

    In external beam radiotherapy, one of the most common and reliable methods for patient geometrical setup and/or predicting the tumor location is use of external markers. In this study, the main challenging issue is increasing the accuracy of patient setup by investigating external markers location. Since the location of each external marker may yield different patient setup accuracy, it is important to assess different locations of external markers using appropriate selective algorithms. To do this, two commercially available algorithms entitled a) canonical correlation analysis (CCA) and b) principal component analysis (PCA) were proposed as input selection algorithms. They work on the basis of maximum correlation coefficient and minimum variance between given datasets. The proposed input selection algorithms work in combination with an adaptive neuro‐fuzzy inference system (ANFIS) as a correlation model to give patient positioning information as output. Our proposed algorithms provide input file of ANFIS correlation model accurately. The required dataset for this study was prepared by means of a NURBS‐based 4D XCAT anthropomorphic phantom that can model the shape and structure of complex organs in human body along with motion information of dynamic organs. Moreover, a database of four real patients undergoing radiation therapy for lung cancers was utilized in this study for validation of proposed strategy. Final analyzed results demonstrate that input selection algorithms can reasonably select specific external markers from those areas of the thorax region where root mean square error (RMSE) of ANFIS model has minimum values at that given area. It is also found that the selected marker locations lie closely in those areas where surface point motion has a large amplitude and a high correlation. PACS number(s): 87.55.km, 87.55.N PMID:27929479

  4. Setup Variations in Radiotherapy of Anal Cancer: Advantages of Target Volume Reduction Using Image-Guided Radiation Treatment

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chen Yijen, E-mail: yichen@coh.org; Suh, Steve; Nelson, Rebecca A.

    2012-09-01

    Purpose: To define setup variations in the radiation treatment (RT) of anal cancer and to report the advantages of image-guided RT (IGRT) in terms of reduction of target volume and treatment-related side effects. Methods and Materials: Twelve consecutive patients with anal cancer treated by combined chemoradiation by use of helical tomotherapy from March 2007 to November 2008 were selected. With patients immobilized and positioned in place, megavoltage computed tomography (MVCT) scans were performed before each treatment and were automatically registered to planning CT scans. Patients were shifted per the registration data and treated. A total of 365 MVCT scans weremore » analyzed. The primary site received a median dose of 55 Gy. To evaluate the potential dosimetric advantage(s) of IGRT, cases were replanned according to Radiation Therapy Oncology Group 0529, with and without adding recommended setup variations from the current study. Results: Significant setup variations were observed throughout the course of RT. The standard deviations for systematic setup correction in the anterior-posterior (AP), lateral, and superior-inferior (SI) directions and roll rotation were 1.1, 3.6, and 3.2 mm, and 0.3 Degree-Sign , respectively. The average random setup variations were 3.8, 5.5, and 2.9 mm, and 0.5 Degree-Sign , respectively. Without daily IGRT, margins of 4.9, 11.1, and 8.5 mm in the AP, lateral, and SI directions would have been needed to ensure that the planning target volume (PTV) received {>=}95% of the prescribed dose. Conversely, daily IGRT required no extra margins on PTV and resulted in a significant reduction of V15 and V45 of intestine and V10 of pelvic bone marrow. Favorable toxicities were observed, except for acute hematologic toxicity. Conclusions: Daily MVCT scans before each treatment can effectively detect setup variations and thereby reduce PTV margins in the treatment of anal cancer. The use of concurrent chemotherapy and IGRT provided favorable toxicities, except for acute hematologic toxicity.« less

  5. Stopping power and dose calculations with analytical and Monte Carlo methods for protons and prompt gamma range verification

    NASA Astrophysics Data System (ADS)

    Usta, Metin; Tufan, Mustafa Çağatay; Aydın, Güral; Bozkurt, Ahmet

    2018-07-01

    In this study, we have performed the calculations stopping power, depth dose, and range verification for proton beams using dielectric and Bethe-Bloch theories and FLUKA, Geant4 and MCNPX Monte Carlo codes. In the framework, as analytical studies, Drude model was applied for dielectric theory and effective charge approach with Roothaan-Hartree-Fock charge densities was used in Bethe theory. In the simulations different setup parameters were selected to evaluate the performance of three distinct Monte Carlo codes. The lung and breast tissues were investigated are considered to be related to the most common types of cancer throughout the world. The results were compared with each other and the available data in literature. In addition, the obtained results were verified with prompt gamma range data. In both stopping power values and depth-dose distributions, it was found that the Monte Carlo values give better results compared with the analytical ones while the results that agree best with ICRU data in terms of stopping power are those of the effective charge approach between the analytical methods and of the FLUKA code among the MC packages. In the depth dose distributions of the examined tissues, although the Bragg curves for Monte Carlo almost overlap, the analytical ones show significant deviations that become more pronounce with increasing energy. Verifications with the results of prompt gamma photons were attempted for 100-200 MeV protons which are regarded important for proton therapy. The analytical results are within 2%-5% and the Monte Carlo values are within 0%-2% as compared with those of the prompt gammas.

  6. Clinical Positioning Accuracy for Multisession Stereotactic Radiotherapy With the Gamma Knife Perfexion

    PubMed Central

    Young, Lori A.; Phillips, Mark H.; Cheung, Michael; Halasz, Lia M.; Rockhill, Jason K.

    2017-01-01

    Multisession stereotactic radiation therapy is increasingly being seen as a preferred option for intracranial diseases in close proximity to critical structures and for larger target volumes. The objective of this study is to investigate the reproducibility of the Extend system from Elekta. A retrospective review was conducted for all patients treated with multisession Gamma Knife between July 2010 and June 2015, including both malignant and benign lesions. Eighty-four patients were treated in this 5-year span. The average residual daily setup uncertainty was 0.48 (0.19) mm. We compare measurements of setup uncertainty from the Extend system to measurements performed with a linac-based approach previously used in our center. The Extend system has significantly reduced setup uncertainty for fractionated intracranial treatments at our institution. Positive results were observed in a small population of edentulous patients. The Extend system compares favorably with other approaches to delivering intracranial stereotactic radiotherapy and is a robust, simple-to-use, and precise method for treating multisession intracranial lesions. PMID:28514899

  7. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zhao, B; Maquilan, G; Anders, M

    Purpose: Full face and neck thermoplastic masks provide standard-of-care immobilization for patients receiving H&N IMRT. However, these masks are uncomfortable and increase skin dose. The purpose of this pilot study was to investigate the feasibility and setup accuracy of open face and neck mask immobilization with OIG. Methods: Ten patients were consented and enrolled to this IRB-approved protocol. Patients were immobilized with open masks securing only forehead and chin. Standard IMRT to 60–70 Gy in 30 fractions were delivered in all cases. Patient simulation information, including isocenter location and CT skin contours, were imported to a commercial OIG system. Onmore » the first day of treatment, patients were initially set up to surface markings and then OIG referenced to face and neck skin regions of interest (ROI) localized on simulation CT images, followed by in-room CBCT. CBCTs were acquired at least weekly while planar OBI was acquired on the days without CBCT. Following 6D robotic couch correction with kV imaging, a new optical real-time surface image was acquired to track intrafraction motion and to serve as a reference surface for setup at the next treatment fraction. Therapists manually recorded total treatment time as well as couch shifts based on kV imaging. Intrafractional ROI motion tracking was automatically recorded. Results: Setup accuracy of OIG was compared with CBCT results. The setup error based on OIG was represented as a 6D shift (vertical/longitudinal/lateral/rotation/pitch/roll). Mean error values were −0.70±3.04mm, −0.69±2.77mm, 0.33±2.67 mm, −0.14±0.94 o, −0.15±1.10o and 0.12±0.82o, respectively for the cohort. Average treatment time was 24.1±9.2 minutes, comparable to standard immobilization. The amplitude of intrafractional ROI motion was 0.69±0.36 mm, driven primarily by respiratory neck motion. Conclusion: OGI can potentially provide accurate setup and treatment tracking for open face and neck immobilization. Study accrual and patient/provider satisfaction survey collection remain ongoing. This study is supported by VisionRT, Ltd.« less

  8. MO-FG-CAMPUS-JeP3-02: A Novel Setup Approach to Improve C-Spine Curvature Reproducibility for Head and Neck Radiotherapy Using Optical Surface Imaging with Two Regions of Interest

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ryan, K; Gil, M; Li, G

    Purpose: To develop a novel approach to improve cervical spine (c-spine) curvature reproducibility for head and neck (HN) patients using optical surface imaging (OSI) with two regions of interests (ROIs). Methods: The OSI-guided, two-step setup procedure requires two ROIs: ROI-1 of the shoulders and ROI-2 of the face. The neck can be stretched or squeezed in superior-inferior (SI) direction using a specially-designed sliding head support. We hypothesize that when these two ROIs are aligned, the c-spine should fall into a naturally reproducible position under same setup conditions. An anthropomorphous phantom test was performed to examine neck pitch angles comparing withmore » the calculated angles. Three volunteers participated in the experiments, which start with conventional HN setup using skin markers and room lasers. An OSI image and lateral photo-picture were acquired as the references. In each of the three replicate tests, conventional setup was first applied after volunteers got on the couch. ROI-1 was aligned by moving the body, followed by ROI-2 alignment via adjusting head position and orientation under real-time OSI guidance. A final static OSI image and lateral picture were taken to evaluate both anterior and posterior surface alignments. Three degrees of freedom can be adjusted if an open-face mask was applied, including head SI shift using the sliding head support and pitch-and-roll rotations using a commercial couch extension. Surface alignment was analyzed comparing with conventional setup. Results: The neck pitch angle measured by OSI is consistent with the calculated (0.2±0.6°). Volunteer study illustrated improved c-spine setup reproducibility using OSI comparing with conventional setup. ROI alignments with 2mm/1° tolerance are achieved within 3 minutes. Identical knee support is important to achieve ROI-1 pitch alignment. Conclusion: The feasibility of this novel approach has been demonstrated for c-spine curvature setup reproducibility. Further evaluation is necessary with bony alignment variation in patient studies. This study is in part supported by the NIH (U54CA137788).« less

  9. Fractionated Proton Radiotherapy for Benign Cavernous Sinus Meningiomas

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Slater, Jerry D., E-mail: jdslater@dominion.llumc.edu; Loredo, Lilia N.; Chung, Arthur

    2012-08-01

    Purpose: To evaluate the efficacy of fractionated proton radiotherapy for a population of patients with benign cavernous sinus meningiomas. Methods and Materials: Between 1991 and 2002, 72 patients were treated at Loma Linda University Medical Center with proton therapy for cavernous sinus meningiomas. Fifty-one patients had biopsy or subtotal resection; 47 had World Health Organization grade 1 pathology. Twenty-one patients had no histologic verification. Twenty-two patients received primary proton therapy; 30 had 1 previous surgery; 20 had more than 1 surgery. The mean gross tumor volume was 27.6 cm{sup 3}; mean clinical target volume was 52.9 cm{sup 3}. Median totalmore » doses for patients with and without histologic verification were 59 and 57 Gy, respectively. Mean and median follow-up periods were 74 months. Results: The overall 5-year actuarial control rate was 96%; the control rate was 99% in patients with grade 1 or absent histologic findings and 50% for those with atypical histology. All 21 patients who did not have histologic verification and 46 of 47 patients with histologic confirmation of grade 1 tumor demonstrated disease control at 5 years. Control rates for patients without previous surgery, 1 surgery, and 2 or more surgeries were 95%, 96%, and 95%, respectively. Conclusions: Fractionated proton radiotherapy for grade 1 cavernous sinus meningiomas achieves excellent control rates with minimal toxicities, regardless of surgical intervention or use of histologic diagnosis. Disease control for large lesions can be achieved by primary fractionated proton therapy.« less

  10. Coupled Facility/Payload Vibration Modeling Improvements

    NASA Technical Reports Server (NTRS)

    Carnahan, Timothy M.; Kaiser, Michael

    2015-01-01

    A major phase of aerospace hardware verification is vibration testing. The standard approach for such testing is to use a shaker to induce loads into the payload. In preparation for vibration testing at NASA/GSFC there is an analysis to assess the responses of the payload. A new method of modeling the test is presented that takes into account dynamic interactions between the facility and the payload. This dynamic interaction has affected testing in the past, but been ignored or adjusted for during testing. By modeling the combination of the facility and test article (payload) it is possible to improve the prediction of hardware responses. Many aerospace test facilities work in similar way to those at NASA Goddard Space Flight Center. Lessons learned here should be applicable to other test facilities with similar setups.

  11. Image Guidance in Radiation Therapy: Techniques and Applications

    PubMed Central

    Kataria, Tejinder

    2014-01-01

    In modern day radiotherapy, the emphasis on reduction on volume exposed to high radiotherapy doses, improving treatment precision as well as reducing radiation-related normal tissue toxicity has increased, and thus there is greater importance given to accurate position verification and correction before delivering radiotherapy. At present, several techniques that accomplish these goals impeccably have been developed, though all of them have their limitations. There is no single method available that eliminates treatment-related uncertainties without considerably adding to the cost. However, delivering “high precision radiotherapy” without periodic image guidance would do more harm than treating large volumes to compensate for setup errors. In the present review, we discuss the concept of image guidance in radiotherapy, the current techniques available, and their expected benefits and pitfalls. PMID:25587445

  12. Application of computer vision to automatic prescription verification in pharmaceutical mail order

    NASA Astrophysics Data System (ADS)

    Alouani, Ali T.

    2005-05-01

    In large volume pharmaceutical mail order, before shipping out prescriptions, licensed pharmacists ensure that the drug in the bottle matches the information provided in the patient prescription. Typically, the pharmacist has about 2 sec to complete the prescription verification process of one prescription. Performing about 1800 prescription verification per hour is tedious and can generate human errors as a result of visual and brain fatigue. Available automatic drug verification systems are limited to a single pill at a time. This is not suitable for large volume pharmaceutical mail order, where a prescription can have as many as 60 pills and where thousands of prescriptions are filled every day. In an attempt to reduce human fatigue, cost, and limit human error, the automatic prescription verification system (APVS) was invented to meet the need of large scale pharmaceutical mail order. This paper deals with the design and implementation of the first prototype online automatic prescription verification machine to perform the same task currently done by a pharmacist. The emphasis here is on the visual aspects of the machine. The system has been successfully tested on 43,000 prescriptions.

  13. SU-F-T-519: Is Geometry Based Setup Sufficient for All of the Head and Neck Treatment Cases?: A Feasibility Study Towards the Dose Based Setup

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lee, S; Chen, S; Zhang, B

    Purpose: This study compares the geometric-based setup (GBS) which is currently used in the clinic to a novel concept of dose-based setup (DBS) of head and neck (H&N) patients using cone beam CT (CBCT) of the day; and evaluates the clinical advantages. Methods: Ten H&N patients who underwent re-simulation and re-plan due to noticeable anatomic changes during the course of the treatments were retrospectively reviewed on dosimetric changes in the assumption of no plan modification was performed. RayStation planning system (RaySearch Laboratories AB, Sweden) was used to match (ROI fusion module) between prescribed isodoseline (IDL) in the CBCT imported alongmore » with ROIs from re-planned CT and the IDL of original plan (Dose-based setup: DBS). Then, the CBCT plan based on daily setup using the GBS (previously used for a patient) and the DBS CBCT plan recalculated in RayStation compared against the original CT-sim plan. Results: Most of patients’ tumor coverage and OAR doses got generally worsen when the CBCT plans were compared with original CT-sim plan with GBS. However, when DBS intervened, the OAR dose and tumor coverage was better than the GBS. For example, one of patients’ daily average doses of right parotid and oral cavity increased to 26% and 36%, respectively from the original plan to the GBS planning. However, it only increased by 13% and 24%, respectively with DBS. GTV D95 coverage also decreased by 16% with GBS, but only 2% decreased with DBS. Conclusion: DBS method is superior to GBS to prevent any abrupt dose changes to OARs as well as PTV/CTV or GTV at least for some H&N cases. Since it is not known when the DBS is beneficial to the GBS, a system which enables the on-line DBS may be helpful for better treatment of H&N.« less

  14. SU-E-J-258: Inter- and Intra-Fraction Setup Stability and Couch Change Tolerance for Image Guided Radiation Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Teboh, Forbang R; Agee, M; Rowe, L

    2014-06-01

    Purpose: Immobilization devices combine rigid patient fixation as well as comfort and play a key role providing the stability required for accurate radiation delivery. In the setup step, couch re-positioning needed to align the patient is derived via registration of acquired versus reference image. For subsequent fractions, replicating the initial setup should yield identical alignment errors when compared to the reference. This is not always the case and further couch re-positioning can be needed. An important quality assurance measure is to set couch tolerances beyond which additional investigations are needed. The purpose of this work was to study the inter-fractionmore » couch changes needed to re-align the patient and the intra-fraction stability of the alignment as a guide to establish the couch tolerances. Methods: Data from twelve patients treated on the Accuray CyberKnife (CK) system for fractionated intracranial radiotherapy and immobilized with Aquaplast RT, U-frame, F-Head-Support (Qfix, PA, USA) was used. Each fraction involved image acquisitions and registration with the reference to re-align the patient. The absolute couch position corresponding to the approved setup alignment was recorded per fraction. Intra-fraction set-up corrections were recorded throughout the treatment. Results: The average approved setup alignment was 0.03±0.28mm, 0.15±0.22mm, 0.06±0.31mm in the L/R, A/P, S/I directions respectively and 0.00±0.35degrees, 0.03±0.32degrees, 0.08±0.45degrees for roll, pitch and yaw respectively. The inter-fraction reproducibility of the couch position was 6.65mm, 10.55mm, and 4.77mm in the L/R, A/P and S/I directions respectively and 0.82degrees, 0.71degrees for roll and pitch respectively. Intra-fraction monitoring showed small average errors of 0.21±0.21mm, 0.00±0.08mm, 0.23±0.22mm in the L/R, A/P, S/I directions respectively and 0.03±0.12degrees, 0.04±0.25degrees, and 0.13±0.15degrees in the roll, pitch and yaw respectively. Conclusion: The inter-fraction reproducibility should serve as a guide to couch tolerances, specific to a site and immobilization. More patients need to be included to make general conclusions.« less

  15. Clinical Experience and Evaluation of Patient Treatment Verification With a Transit Dosimeter

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ricketts, Kate, E-mail: k.ricketts@ucl.ac.uk; Department of Radiotherapy Physics, Royal Berkshire NHS Foundation Trust, Reading; Navarro, Clara

    2016-08-01

    Purpose: To prospectively evaluate a protocol for transit dosimetry on a patient population undergoing intensity modulated radiation therapy (IMRT) and to assess the issues in clinical implementation of electronic portal imaging devices (EPIDs) for treatment verification. Methods and Materials: Fifty-eight patients were enrolled in the study. Amorphous silicon EPIDs were calibrated for dose and used to acquire images of delivered fields. Measured EPID dose maps were back-projected using the planning computed tomographic (CT) images to calculate dose at prespecified points within the patient and compared with treatment planning system dose offline using point dose difference and point γ analysis. Themore » deviation of the results was used to inform future action levels. Results: Two hundred twenty-five transit images were analyzed, composed of breast, prostate, and head and neck IMRT fields. Patient measurements demonstrated the potential of the dose verification protocol to model dose well under complex conditions: 83.8% of all delivered beams achieved the initial set tolerance level of Δ{sub D} of 0 ± 5 cGy or %Δ{sub D} of 0% ± 5%. Importantly, the protocol was also sensitive to anatomic changes and spotted that 3 patients from 20 measured prostate patients had undergone anatomic change in comparison with the planning CT. Patient data suggested an EPID-reconstructed versus treatment planning system dose difference action level of 0% ± 7% for breast fields. Asymmetric action levels were more appropriate for inversed IMRT fields, using absolute dose difference (−2 ± 5 cGy) or summed field percentage dose difference (−6% ± 7%). Conclusions: The in vivo dose verification method was easy to use and simple to implement, and it could detect patient anatomic changes that impacted dose delivery. The system required no extra dose to the patient or treatment time delay and so could be used throughout the course of treatment to identify and limit systematic and random errors in dose delivery for patient groups.« less

  16. Clinical Experience and Evaluation of Patient Treatment Verification With a Transit Dosimeter.

    PubMed

    Ricketts, Kate; Navarro, Clara; Lane, Katherine; Blowfield, Claire; Cotten, Gary; Tomala, Dee; Lord, Christine; Jones, Joanne; Adeyemi, Abiodun

    2016-08-01

    To prospectively evaluate a protocol for transit dosimetry on a patient population undergoing intensity modulated radiation therapy (IMRT) and to assess the issues in clinical implementation of electronic portal imaging devices (EPIDs) for treatment verification. Fifty-eight patients were enrolled in the study. Amorphous silicon EPIDs were calibrated for dose and used to acquire images of delivered fields. Measured EPID dose maps were back-projected using the planning computed tomographic (CT) images to calculate dose at prespecified points within the patient and compared with treatment planning system dose offline using point dose difference and point γ analysis. The deviation of the results was used to inform future action levels. Two hundred twenty-five transit images were analyzed, composed of breast, prostate, and head and neck IMRT fields. Patient measurements demonstrated the potential of the dose verification protocol to model dose well under complex conditions: 83.8% of all delivered beams achieved the initial set tolerance level of ΔD of 0 ± 5 cGy or %ΔD of 0% ± 5%. Importantly, the protocol was also sensitive to anatomic changes and spotted that 3 patients from 20 measured prostate patients had undergone anatomic change in comparison with the planning CT. Patient data suggested an EPID-reconstructed versus treatment planning system dose difference action level of 0% ± 7% for breast fields. Asymmetric action levels were more appropriate for inversed IMRT fields, using absolute dose difference (-2 ± 5 cGy) or summed field percentage dose difference (-6% ± 7%). The in vivo dose verification method was easy to use and simple to implement, and it could detect patient anatomic changes that impacted dose delivery. The system required no extra dose to the patient or treatment time delay and so could be used throughout the course of treatment to identify and limit systematic and random errors in dose delivery for patient groups. Copyright © 2016 Elsevier Inc. All rights reserved.

  17. Total skin electron therapy in the lying‐on‐the‐floor position using a customized flattening filter to accommodate frail patients

    PubMed Central

    Antolak, John A.

    2013-01-01

    A total skin electron (TSE) floor technique is presented for treating patients who are unable to safely stand for extended durations. A customized flattening filter is used to eliminate the need for field junctioning, improve field uniformity, and reduce setup time. The flattening filter is constructed from copper and polycarbonate, fits into the linac's accessory slot, and is optimized to extend the useful height and width of the beam such that no field junctions are needed during treatment. A TSE floor with flattening filter (TSE FF) treatment course consisted of six patient positions: three supine and three prone. For all treatment fields, electron beam energy was 6 MeV; collimator settings were an x of 30 cm, y of 40 cm, and θcoll of 0°; and a 0.4 cm thick polycarbonate spoiler was positioned in front of the patient. Percent depth dose (PDD) and photon contamination for the TSE FF technique were compared with our standard technique, which is similar to the Stanford technique. Beam profiles were measured using radiochromic film, and dose uniformity was verified using an anthropomorphic radiological phantom. The TSE FF technique met field uniformity requirements specified by the American Association of Physicists in Medicine Task Group 30. TSE FF R80 ranges from 4 to 4.8 mm. TSE FF photon contamination was ~ 3%. Anthropomorphic radiological phantom verification demonstrated that dose to the entire skin surface was expected to be within about ±15% of the prescription dose, except for the perineum, scalp vertex, top of shoulder, and soles of the feet. The TSE floor technique presented herein eliminates field junctioning, is suitable for patients who cannot safely stand during treatment, and provides comparable quality and uniformity to the Stanford technique. PACS number: 87 PMID:24036864

  18. Kilovoltage cone-beam CT imaging dose during breast radiotherapy: A dose comparison between a left and right breast setup

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Quinn, Alexandra, E-mail: Alexandra.quinn@health.nsw.gov.au; Centre for Medical Radiation Physics, University of Wollongong, NSW; Liverpool and Macarthur Cancer Therapy Centres, NSW

    2014-07-01

    The purpose of this study was to investigate the delivered dose from a kilovoltage cone-beam computed tomography (kV-CBCT) acquired in breast treatment position for a left and right breast setup. The dose was measured with thermoluminescent dosimeters positioned within a female anthropomorphic phantom at organ locations. Imaging was performed on an Elekta Synergy XVI system with the phantom setup on a breast board. The image protocol involved 120 kVp, 140 mAs, and a 270° arc rotation clockwise 0° to 270° for the left breast setup and 270° to 180° for the right breast setup (maximum arc rotations possible). The dosemore » delivered to the left breast, right breast, and heart was 5.1 mGy, 3.9 mGy, and 4.0 mGy for the left breast setup kV-CBCT, and 6.4 mGy, 6.0 mGy, and 4.8 mGy for the right breast setup kV-CBCT, respectively. The rotation arc of the kV-CBCT influenced the dose delivered, with the right breast setup kV-CBCT found to deliver a dose of up to 4 mGy or 105% higher to the treated breast′s surface in comparison with the left breast setup. This is attributed to the kV-CBCT source being more proximal to the anterior of the phantom for a right breast setup, whereas the source is more proximal to the posterior of the patient for a left-side scan.« less

  19. Spot scanning proton therapy plan assessment: design and development of a dose verification application for use in routine clinical practice

    NASA Astrophysics Data System (ADS)

    Augustine, Kurt E.; Walsh, Timothy J.; Beltran, Chris J.; Stoker, Joshua B.; Mundy, Daniel W.; Parry, Mark D.; Bues, Martin; Fatyga, Mirek

    2016-04-01

    The use of radiation therapy for the treatment of cancer has been carried out clinically since the late 1800's. Early on however, it was discovered that a radiation dose sufficient to destroy cancer cells can also cause severe injury to surrounding healthy tissue. Radiation oncologists continually strive to find the perfect balance between a dose high enough to destroy the cancer and one that avoids damage to healthy organs. Spot scanning or "pencil beam" proton radiotherapy offers another option to improve on this. Unlike traditional photon therapy, proton beams stop in the target tissue, thus better sparing all organs beyond the targeted tumor. In addition, the beams are far narrower and thus can be more precisely "painted" onto the tumor, avoiding exposure to surrounding healthy tissue. To safely treat patients with proton beam radiotherapy, dose verification should be carried out for each plan prior to treatment. Proton dose verification systems are not currently commercially available so the Department of Radiation Oncology at the Mayo Clinic developed its own, called DOSeCHECK, which offers two distinct dose simulation methods: GPU-based Monte Carlo and CPU-based analytical. The three major components of the system include the web-based user interface, the Linux-based dose verification simulation engines, and the supporting services and components. The architecture integrates multiple applications, libraries, platforms, programming languages, and communication protocols and was successfully deployed in time for Mayo Clinic's first proton beam therapy patient. Having a simple, efficient application for dose verification greatly reduces staff workload and provides additional quality assurance, ultimately improving patient safety.

  20. Comparison between In-house developed and Diamond commercial software for patient specific independent monitor unit calculation and verification with heterogeneity corrections.

    PubMed

    Kuppusamy, Vijayalakshmi; Nagarajan, Vivekanandan; Jeevanandam, Prakash; Murugan, Lavanya

    2016-02-01

    The study was aimed to compare two different monitor unit (MU) or dose verification software in volumetric modulated arc therapy (VMAT) using modified Clarkson's integration technique for 6 MV photons beams. In-house Excel Spreadsheet based monitor unit verification calculation (MUVC) program and PTW's DIAMOND secondary check software (SCS), version-6 were used as a secondary check to verify the monitor unit (MU) or dose calculated by treatment planning system (TPS). In this study 180 patients were grouped into 61 head and neck, 39 thorax and 80 pelvic sites. Verification plans are created using PTW OCTAVIUS-4D phantom and also measured using 729 detector chamber and array with isocentre as the suitable point of measurement for each field. In the analysis of 154 clinically approved VMAT plans with isocentre at a region above -350 HU, using heterogeneity corrections, In-house Spreadsheet based MUVC program and Diamond SCS showed good agreement TPS. The overall percentage average deviations for all sites were (-0.93% + 1.59%) and (1.37% + 2.72%) for In-house Excel Spreadsheet based MUVC program and Diamond SCS respectively. For 26 clinically approved VMAT plans with isocentre at a region below -350 HU showed higher variations for both In-house Spreadsheet based MUVC program and Diamond SCS. It can be concluded that for patient specific quality assurance (QA), the In-house Excel Spreadsheet based MUVC program and Diamond SCS can be used as a simple and fast accompanying to measurement based verification for plans with isocentre at a region above -350 HU. Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  1. Personal Verification/Identification via Analysis of the Peripheral ECG Leads: Influence of the Personal Health Status on the Accuracy

    PubMed Central

    Bortolan, Giovanni

    2015-01-01

    Traditional means for identity validation (PIN codes, passwords), and physiological and behavioral biometric characteristics (fingerprint, iris, and speech) are susceptible to hacker attacks and/or falsification. This paper presents a method for person verification/identification based on correlation of present-to-previous limb ECG leads: I (r I), II (r II), calculated from them first principal ECG component (r PCA), linear and nonlinear combinations between r I, r II, and r PCA. For the verification task, the one-to-one scenario is applied and threshold values for r I, r II, and r PCA and their combinations are derived. The identification task supposes one-to-many scenario and the tested subject is identified according to the maximal correlation with a previously recorded ECG in a database. The population based ECG-ILSA database of 540 patients (147 healthy subjects, 175 patients with cardiac diseases, and 218 with hypertension) has been considered. In addition a common reference PTB dataset (14 healthy individuals) with short time interval between the two acquisitions has been taken into account. The results on ECG-ILSA database were satisfactory with healthy people, and there was not a significant decrease in nonhealthy patients, demonstrating the robustness of the proposed method. With PTB database, the method provides an identification accuracy of 92.9% and a verification sensitivity and specificity of 100% and 89.9%. PMID:26568954

  2. Personal Verification/Identification via Analysis of the Peripheral ECG Leads: Influence of the Personal Health Status on the Accuracy.

    PubMed

    Jekova, Irena; Bortolan, Giovanni

    2015-01-01

    Traditional means for identity validation (PIN codes, passwords), and physiological and behavioral biometric characteristics (fingerprint, iris, and speech) are susceptible to hacker attacks and/or falsification. This paper presents a method for person verification/identification based on correlation of present-to-previous limb ECG leads: I (r I), II (r II), calculated from them first principal ECG component (r PCA), linear and nonlinear combinations between r I, r II, and r PCA. For the verification task, the one-to-one scenario is applied and threshold values for r I, r II, and r PCA and their combinations are derived. The identification task supposes one-to-many scenario and the tested subject is identified according to the maximal correlation with a previously recorded ECG in a database. The population based ECG-ILSA database of 540 patients (147 healthy subjects, 175 patients with cardiac diseases, and 218 with hypertension) has been considered. In addition a common reference PTB dataset (14 healthy individuals) with short time interval between the two acquisitions has been taken into account. The results on ECG-ILSA database were satisfactory with healthy people, and there was not a significant decrease in nonhealthy patients, demonstrating the robustness of the proposed method. With PTB database, the method provides an identification accuracy of 92.9% and a verification sensitivity and specificity of 100% and 89.9%.

  3. Simulated Order Verification and Medication Reconciliation during an Introductory Pharmacy Practice Experience.

    PubMed

    Metzger, Nicole L; Chesson, Melissa M; Momary, Kathryn M

    2015-09-25

    Objective. To create, implement, and assess a simulated medication reconciliation and an order verification activity using hospital training software. Design. A simulated patient with medication orders and home medications was built into existing hospital training software. Students in an institutional introductory pharmacy practice experience (IPPE) reconciled the patient's medications and determined whether or not to verify the inpatient orders based on his medical history and laboratory data. After reconciliation, students identified medication discrepancies and documented their rationale for rejecting inpatient orders. Assessment. For a 3-year period, the majority of students agreed the simulation enhanced their learning, taught valuable clinical decision-making skills, integrated material from previous courses, and stimulated their interest in institutional pharmacy. Overall feedback from student evaluations about the IPPE also was favorable. Conclusion. Use of existing hospital training software can affordably simulate the pharmacist's role in order verification and medication reconciliation, as well as improve clinical decision-making.

  4. Frameless fractionated stereotactic radiation therapy of intracranial lesions: impact of cone beam CT based setup correction on dose distribution

    PubMed Central

    2013-01-01

    Background The purpose of this study was to evaluate the impact of Cone Beam CT (CBCT) based setup correction on total dose distributions in fractionated frameless stereotactic radiation therapy of intracranial lesions. Methods Ten patients with intracranial lesions treated with 30 Gy in 6 fractions were included in this study. Treatment planning was performed with Oncentra® for a SynergyS® (Elekta Ltd, Crawley, UK) linear accelerator with XVI® Cone Beam CT, and HexaPOD™ couch top. Patients were immobilized by thermoplastic masks (BrainLab, Reuther). After initial patient setup with respect to lasers, a CBCT study was acquired and registered to the planning CT (PL-CT) study. Patient positioning was corrected according to the correction values (translational, rotational) calculated by the XVI® system. Afterwards a second CBCT study was acquired and registered to the PL-CT to confirm the accuracy of the corrections. An in-house developed software was used for rigid transformation of the PL-CT to the CBCT geometry, and dose calculations for each fraction were performed on the transformed CT. The total dose distribution was achieved by back-transformation and summation of the dose distributions of each fraction. Dose distributions based on PL-CT, CBCT (laser set-up), and final CBCT were compared to assess the influence of setup inaccuracies. Results The mean displacement vector, calculated over all treatments, was reduced from (4.3 ± 1.3) mm for laser based setup to (0.5 ± 0.2) mm if CBCT corrections were applied. The mean rotational errors around the medial-lateral, superior-inferior, anterior-posterior axis were reduced from (−0.1 ± 1.4)°, (0.1 ± 1.2)° and (−0.2 ± 1.0)°, to (0.04 ± 0.4)°, (0.01 ± 0.4)° and (0.02 ± 0.3)°. As a consequence the mean deviation between planned and delivered dose in the planning target volume (PTV) could be reduced from 12.3% to 0.4% for D95 and from 5.9% to 0.1% for Dav. Maximum deviation was reduced from 31.8% to 0.8% for D95, and from 20.4% to 0.1% for Dav. Conclusion Real dose distributions differ substantially from planned dose distributions, if setup is performed according to lasers only. Thermoplasic masks combined with a daily CBCT enabled a sufficient accuracy in dose distribution. PMID:23800172

  5. Radiation Hardness of dSiPM Sensors in a Proton Therapy Radiation Environment

    NASA Astrophysics Data System (ADS)

    Diblen, Faruk; Buitenhuis, Tom; Solf, Torsten; Rodrigues, Pedro; van der Graaf, Emiel; van Goethem, Marc-Jan; Brandenburg, Sytze; Dendooven, Peter

    2017-07-01

    In vivo verification of dose delivery in proton therapy by means of positron emission tomography (PET) or prompt gamma imaging is mostly based on fast scintillation detectors. The digital silicon photomultiplier (dSiPM) allows excellent scintillation detector timing properties and is thus being considered for such verification methods. We present here the results of the first investigation of radiation damage to dSiPM sensors in a proton therapy radiation environment. Radiation hardness experiments were performed at the AGOR cyclotron facility at the KVI-Center for Advanced Radiation Technology, University of Groningen. A 150-MeV proton beam was fully stopped in a water target. In the first experiment, bare dSiPM sensors were placed at 25 cm from the Bragg peak, perpendicular to the beam direction, a geometry typical for an in situ implementation of a PET or prompt gamma imaging device. In the second experiment, dSiPM-based PET detectors containing lutetium yttrium orthosilicate scintillator crystal arrays were placed at 2 and 4 m from the Bragg peak, perpendicular to the beam direction; resembling an in-room PET implementation. Furthermore, the experimental setup was simulated with a Geant4-based Monte Carlo code in order to determine the angular and energy distributions of the neutrons and to determine the 1-MeV equivalent neutron fluences delivered to the dSiPM sensors. A noticeable increase in dark count rate (DCR) after an irradiation with about 108 1-MeV equivalent neutrons/cm2 agrees with observations by others for analog SiPMs, indicating that the radiation damage occurs in the single photon avalanche diodes and not in the electronics integrated on the sensor chip. It was found that in the in situ location, the DCR becomes too large for successful operation after the equivalent of a few weeks of use in a proton therapy treatment room (about 5 × 1013 protons). For PET detectors in an in-room setup, detector performance was unchanged even after an irradiation equivalent to three years of use in a treatment room (3 × 1015 protons).

  6. A procedure and program to calculate shuttle mask advantage

    NASA Astrophysics Data System (ADS)

    Balasinski, A.; Cetin, J.; Kahng, A.; Xu, X.

    2006-10-01

    A well-known recipe for reducing mask cost component in product development is to place non-redundant elements of layout databases related to multiple products on one reticle plate [1,2]. Such reticles are known as multi-product, multi-layer, or, in general, multi-IP masks. The composition of the mask set should minimize not only the layout placement cost, but also the cost of the manufacturing process, design flow setup, and product design and introduction to market. An important factor is the quality check which should be expeditious and enable thorough visual verification to avoid costly modifications once the data is transferred to the mask shop. In this work, in order to enable the layer placement and quality check procedure, we proposed an algorithm where mask layers are first lined up according to the price and field tone [3]. Then, depending on the product die size, expected fab throughput, and scribeline requirements, the subsequent product layers are placed on the masks with different grades. The actual reduction of this concept to practice allowed us to understand the tradeoffs between the automation of layer placement and setup related constraints. For example, the limited options of the numbers of layer per plate dictated by the die size and other design feedback, made us consider layer pairing based not only on the final price of the mask set, but also on the cost of mask design and fab-friendliness. We showed that it may be advantageous to introduce manual layer pairing to ensure that, e.g., all interconnect layers would be placed on the same plate, allowing for easy and simultaneous design fixes. Another enhancement was to allow some flexibility in mixing and matching of the layers such that non-critical ones requiring low mask grade would be placed in a less restrictive way, to reduce the count of orphan layers. In summary, we created a program to automatically propose and visualize shuttle mask architecture for design verification, with enhancements to due to the actual application of the code.

  7. A randomized control trial evaluating fluorescent ink versus dark ink tattoos for breast radiotherapy

    PubMed Central

    Kirby, Anna M; Lee, Steven F; Bartlett, Freddie; Titmarsh, Kumud; Donovan, Ellen; Griffin, Clare L; Gothard, Lone; Locke, Imogen; McNair, Helen A

    2016-01-01

    Objective: The purpose of this UK study was to evaluate interfraction reproducibility and body image score when using ultraviolet (UV) tattoos (not visible in ambient lighting) for external references during breast/chest wall radiotherapy and compare with conventional dark ink. Methods: In this non-blinded, single-centre, parallel group, randomized control trial, patients were allocated to receive either conventional dark ink or UV ink tattoos using computer-generated random blocks. Participant assignment was not masked. Systematic (∑) and random (σ) setup errors were determined using electronic portal images. Body image questionnaires were completed at pre-treatment, 1 month and 6 months to determine the impact of tattoo type on body image. The primary end point was to determine that UV tattoo random error (σsetup) was no less accurate than with conventional dark ink tattoos, i.e. <2.8 mm. Results: 46 patients were randomized to receive conventional dark or UV ink tattoos. 45 patients completed treatment (UV: n = 23, dark: n = 22). σsetup for the UV tattoo group was <2.8 mm in the u and v directions (p = 0.001 and p = 0.009, respectively). A larger proportion of patients reported improvement in body image score in the UV tattoo group compared with the dark ink group at 1 month [56% (13/23) vs 14% (3/22), respectively] and 6 months [52% (11/21) vs 38% (8/21), respectively]. Conclusion: UV tattoos were associated with interfraction setup reproducibility comparable with conventional dark ink. Patients reported a more favourable change in body image score up to 6 months following treatment. Advances in knowledge: This study is the first to evaluate UV tattoo external references in a randomized control trial. PMID:27710100

  8. A randomized control trial evaluating fluorescent ink versus dark ink tattoos for breast radiotherapy.

    PubMed

    Landeg, Steven J; Kirby, Anna M; Lee, Steven F; Bartlett, Freddie; Titmarsh, Kumud; Donovan, Ellen; Griffin, Clare L; Gothard, Lone; Locke, Imogen; McNair, Helen A

    2016-12-01

    The purpose of this UK study was to evaluate interfraction reproducibility and body image score when using ultraviolet (UV) tattoos (not visible in ambient lighting) for external references during breast/chest wall radiotherapy and compare with conventional dark ink. In this non-blinded, single-centre, parallel group, randomized control trial, patients were allocated to receive either conventional dark ink or UV ink tattoos using computer-generated random blocks. Participant assignment was not masked. Systematic (∑) and random (σ) setup errors were determined using electronic portal images. Body image questionnaires were completed at pre-treatment, 1 month and 6 months to determine the impact of tattoo type on body image. The primary end point was to determine that UV tattoo random error (σ setup ) was no less accurate than with conventional dark ink tattoos, i.e. <2.8 mm. 46 patients were randomized to receive conventional dark or UV ink tattoos. 45 patients completed treatment (UV: n = 23, dark: n = 22). σ setup for the UV tattoo group was <2.8 mm in the u and v directions (p = 0.001 and p = 0.009, respectively). A larger proportion of patients reported improvement in body image score in the UV tattoo group compared with the dark ink group at 1 month [56% (13/23) vs 14% (3/22), respectively] and 6 months [52% (11/21) vs 38% (8/21), respectively]. UV tattoos were associated with interfraction setup reproducibility comparable with conventional dark ink. Patients reported a more favourable change in body image score up to 6 months following treatment. Advances in knowledge: This study is the first to evaluate UV tattoo external references in a randomized control trial.

  9. Immobilization precision of a modified GTC frame.

    PubMed

    Winey, Brian; Daartz, Juliane; Dankers, Frank; Bussière, Marc

    2012-05-10

    The purpose of this study was to evaluate and quantify the interfraction reproducibility and intrafraction immobilization precision of a modified GTC frame. The error of the patient alignment and imaging systems were measured using a cranial skull phantom, with simulated, predetermined shifts. The kV setup images were acquired with a room-mounted set of kV sources and panels. Calculated translations and rotations provided by the computer alignment software relying upon three implanted fiducials were compared to the known shifts, and the accuracy of the imaging and positioning systems was calculated. Orthogonal kV setup images for 45 proton SRT patients and 1002 fractions (average 22.3 fractions/patient) were analyzed for interfraction and intrafraction immobilization precision using a modified GTC frame. The modified frame employs a radiotransparent carbon cup and molded pillow to allow for more treatment angles from posterior directions for cranial lesions. Patients and the phantom were aligned with three 1.5 mm stainless steel fiducials implanted into the skull. The accuracy and variance of the patient positioning and imaging systems were measured to be 0.10 ± 0.06 mm, with the maximum uncertainty of rotation being ±0.07°. 957 pairs of interfraction image sets and 974 intrafraction image sets were analyzed. 3D translations and rotations were recorded. The 3D vector interfraction setup reproducibility was 0.13 mm ± 1.8 mm for translations and the largest uncertainty of ± 1.07º for rotations. The intrafraction immobilization efficacy was 0.19 mm ± 0.66 mm for translations and the largest uncertainty of ± 0.50º for rotations. The modified GTC frame provides reproducible setup and effective intrafraction immobilization, while allowing for the complete range of entrance angles from the posterior direction.

  10. Radiation therapy planning and simulation with magnetic resonance images

    NASA Astrophysics Data System (ADS)

    Boettger, Thomas; Nyholm, Tufve; Karlsson, Magnus; Nunna, Chandrasekhar; Celi, Juan Carlos

    2008-03-01

    We present a system which allows for use of magnetic resonance (MR) images as primary RT workflow modality alone and no longer limits the user to computed tomography data for radiation therapy (RT) planning, simulation and patient localization. The single steps for achieving this goal are explained in detail. For planning two MR data sets, MR1 and MR2 are acquired sequentially. For MR1 a standardized Ultrashort TE (UTE) sequence is used enhancing bony anatomy. The sequence for MR2 is chosen to get optimal contrast for the target and the organs at risk for each individual patient. Both images are naturally in registration, neglecting elastic soft tissue deformations. The planning software first automatically extracts skin and bony anatomy from MR1. The user can semi-automatically delineate target structures and organs at risk based on MR1 or MR2, associate all segmentations with MR1 and create a plan in the coordinate system of MR1. Projections similar to digitally reconstructed radiographs (DRR) enhancing bony anatomy are calculated from the MR1 directly and can be used for iso-center definition and setup verification. Furthermore we present a method for creating a Pseudo-CT data set which assigns electron densities to the voxels of MR1 based on the skin and bone segmentations. The Pseudo-CT is then used for dose calculation. Results from first tests under clinical conditions show the feasibility of the completely MR based workflow in RT for necessary clinical cases. It needs to be investigated in how far geometrical distortions influence accuracy of MR-based RT planning.

  11. SU-C-204-06: Surface Imaging for the Set-Up of Proton Post-Mastectomy Chestwall Irradiation: Gated Images Vs Non Gated Images

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Batin, E; Depauw, N; MacDonald, S

    Purpose: Historically, the set-up for proton post-mastectomy chestwall irradiation at our institution started with positioning the patient using tattoos and lasers. One or more rounds of orthogonal X-rays at gantry 0° and beamline X-ray at treatment gantry angle were then taken to finalize the set-up position. As chestwall targets are shallow and superficial, surface imaging is a promising tool for set-up and needs to be investigated Methods: The orthogonal imaging was entirely replaced by AlignRT™ (ART) images. The beamline X-Ray image is kept as a confirmation, based primarily on three opaque markers placed on skin surface instead of bony anatomy.more » In the first phase of the process, ART gated images were used to set-up the patient and the same specific point of the breathing curve was used every day. The moves (translations and rotations) computed for each point of the breathing curve during the first five fractions were analyzed for ten patients. During a second phase of the study, ART gated images were replaced by ART non-gated images combined with real-time monitoring. In both cases, ART images were acquired just before treatment to access the patient position compare to the non-gated CT. Results: The average difference between the maximum move and the minimum move depending on the chosen breathing curve point was less than 1.7 mm for all translations and less than 0.7° for all rotations. The average position discrepancy over the course of treatment obtained by ART non gated images combined to real-time monitoring taken before treatment to the planning CT were smaller than the average position discrepancy obtained using ART gated images. The X-Ray validation images show similar results with both ART imaging process. Conclusion: The use of ART non gated images combined with real time imaging allows positioning post-mastectomy chestwall patients in less than 3 mm / 1°.« less

  12. MO-G-BRE-03: Automated Continuous Monitoring of Patient Setup with Second-Check Independent Image Registration

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jiang, X; Fox, T; Schreibmann, E

    2014-06-15

    Purpose: To create a non-supervised quality assurance program to monitor image-based patient setup. The system acts a secondary check by independently computing shifts and rotations and interfaces with Varian's database to verify therapist's work and warn against sub-optimal setups. Methods: Temporary digitally-reconstructed radiographs (DRRs) and OBI radiographic image files created by Varian's treatment console during patient setup are intercepted and used as input in an independent registration module customized for accuracy that determines the optimal rotations and shifts. To deal with the poor quality of OBI images, a histogram equalization of the live images to the DDR counterparts is performedmore » as a pre-processing step. A search for the most sensitive metric was performed by plotting search spaces subject to various translations and convergence analysis was applied to ensure the optimizer finds the global minima. Final system configuration uses the NCC metric with 150 histogram bins and a one plus one optimizer running for 2000 iterations with customized scales for translations and rotations in a multi-stage optimization setup that first corrects and translations and subsequently rotations. Results: The system was installed clinically to monitor and provide almost real-time feedback on patient positioning. On a 2 month-basis uncorrected pitch values were of a mean 0.016° with standard deviation of 1.692°, and couch rotations of − 0.090°± 1.547°. The couch shifts were −0.157°±0.466° cm for the vertical, 0.045°±0.286 laterally and 0.084°± 0.501° longitudinally. Uncorrected pitch angles were the most common source of discrepancies. Large variations in the pitch angles were correlated with patient motion inside the mask. Conclusion: A system for automated quality assurance of therapist's registration was designed and tested in clinical practice. The approach complements the clinical software's automated registration in terms of algorithm configuration and performance and constitutes a practical approach to implement safe and cost-effective radiotherapy.« less

  13. Maxillary molar derotation and distalization by using a nickel-titanium wire fabricated on a setup model

    PubMed Central

    Jung, Jong Moon; Wi, Young Joo; Koo, Hyun Mo; Kim, Min Ji

    2017-01-01

    The purpose of this article is to introduce a simple appliance that uses a setup model and a nickel-titanium (Ni-Ti) wire for correcting the mesial rotation and drift of the permanent maxillary first molar. The technique involves bonding a Ni-Ti wire to the proper position of the target tooth on a setup model, followed by the fabrication of the transfer cap for indirect bonding and its transfer to the patient's teeth. This appliance causes less discomfort and provides better oral hygiene for the patients than do conventional appliances such as the bracket, pendulum, and distal jet. The treatment time is also shorter with the new appliance than with full-fixed appliances. Moreover, the applicability of the new appliance can be expanded to many cases by using screws or splinting with adjacent teeth to improve anchorage. PMID:28670568

  14. Maxillary molar derotation and distalization by using a nickel-titanium wire fabricated on a setup model.

    PubMed

    Jung, Jong Moon; Wi, Young Joo; Koo, Hyun Mo; Kim, Min Ji; Chun, Youn Sic

    2017-07-01

    The purpose of this article is to introduce a simple appliance that uses a setup model and a nickel-titanium (Ni-Ti) wire for correcting the mesial rotation and drift of the permanent maxillary first molar. The technique involves bonding a Ni-Ti wire to the proper position of the target tooth on a setup model, followed by the fabrication of the transfer cap for indirect bonding and its transfer to the patient's teeth. This appliance causes less discomfort and provides better oral hygiene for the patients than do conventional appliances such as the bracket, pendulum, and distal jet. The treatment time is also shorter with the new appliance than with full-fixed appliances. Moreover, the applicability of the new appliance can be expanded to many cases by using screws or splinting with adjacent teeth to improve anchorage.

  15. MO-FG-202-01: A Fast Yet Sensitive EPID-Based Real-Time Treatment Verification System

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ahmad, M; Nourzadeh, H; Neal, B

    2016-06-15

    Purpose: To create a real-time EPID-based treatment verification system which robustly detects treatment delivery and patient attenuation variations. Methods: Treatment plan DICOM files sent to the record-and-verify system are captured and utilized to predict EPID images for each planned control point using a modified GPU-based digitally reconstructed radiograph algorithm which accounts for the patient attenuation, source energy fluence, source size effects, and MLC attenuation. The DICOM and predicted images are utilized by our C++ treatment verification software which compares EPID acquired 1024×768 resolution frames acquired at ∼8.5hz from Varian Truebeam™ system. To maximize detection sensitivity, image comparisons determine (1) ifmore » radiation exists outside of the desired treatment field; (2) if radiation is lacking inside the treatment field; (3) if translations, rotations, and magnifications of the image are within tolerance. Acquisition was tested with known test fields and prior patient fields. Error detection was tested in real-time and utilizing images acquired during treatment with another system. Results: The computational time of the prediction algorithms, for a patient plan with 350 control points and 60×60×42cm^3 CT volume, is 2–3minutes on CPU and <27 seconds on GPU for 1024×768 images. The verification software requires a maximum of ∼9ms and ∼19ms for 512×384 and 1024×768 resolution images, respectively, to perform image analysis and dosimetric validations. Typical variations in geometric parameters between reference and the measured images are 0.32°for gantry rotation, 1.006 for scaling factor, and 0.67mm for translation. For excess out-of-field/missing in-field fluence, with masks extending 1mm (at isocenter) from the detected aperture edge, the average total in-field area missing EPID fluence was 1.5mm2 the out-of-field excess EPID fluence was 8mm^2, both below error tolerances. Conclusion: A real-time verification software, with EPID images prediction algorithm, was developed. The system is capable of performing verifications between frames acquisitions and identifying source(s) of any out-of-tolerance variations. This work was supported in part by Varian Medical Systems.« less

  16. A silicon strip detector array for energy verification and quality assurance in heavy ion therapy.

    PubMed

    Debrot, Emily; Newall, Matthew; Guatelli, Susanna; Petasecca, Marco; Matsufuji, Naruhiro; Rosenfeld, Anatoly B

    2018-02-01

    The measurement of depth dose profiles for range and energy verification of heavy ion beams is an important aspect of quality assurance procedures for heavy ion therapy facilities. The steep dose gradients in the Bragg peak region of these profiles require the use of detectors with high spatial resolution. The aim of this work is to characterize a one dimensional monolithic silicon detector array called the "serial Dose Magnifying Glass" (sDMG) as an independent ion beam energy and range verification system used for quality assurance conducted for ion beams used in heavy ion therapy. The sDMG detector consists of two linear arrays of 128 silicon sensitive volumes each with an effective size of 2mm × 50μm × 100μm fabricated on a p-type substrate at a pitch of 200 μm along a single axis of detection. The detector was characterized for beam energy and range verification by measuring the response of the detector when irradiated with a 290 MeV/u 12 C ion broad beam incident along the single axis of the detector embedded in a PMMA phantom. The energy of the 12 C ion beam incident on the detector and the residual energy of an ion beam incident on the phantom was determined from the measured Bragg peak position in the sDMG. Ad hoc Monte Carlo simulations of the experimental setup were also performed to give further insight into the detector response. The relative response profiles along the single axis measured with the sDMG detector were found to have good agreement between experiment and simulation with the position of the Bragg peak determined to fall within 0.2 mm or 1.1% of the range in the detector for the two cases. The energy of the beam incident on the detector was found to vary less than 1% between experiment and simulation. The beam energy incident on the phantom was determined to be (280.9 ± 0.8) MeV/u from the experimental and (280.9 ± 0.2) MeV/u from the simulated profiles. These values coincide with the expected energy of 281 MeV/u. The sDMG detector response was studied experimentally and characterized using a Monte Carlo simulation. The sDMG detector was found to accurately determine the 12 C beam energy and is suited for fast energy and range verification quality assurance. It is proposed that the sDMG is also applicable for verification of treatment planning systems that rely on particle range. © 2017 American Association of Physicists in Medicine.

  17. Surgical treatment of gastroesophageal reflux disease and upside-down stomach using the Da Vinci robotic system. A prospective study.

    PubMed

    Hartmann, Jens; Jacobi, Christoph A; Menenakos, Charalambos; Ismail, Mahmoud; Braumann, Chris

    2008-03-01

    So far, the impact of telematic surgical approach in Gastroesophageal Reflux Disease (GERD) is still obscure. In this prospective study, we analyzed the Da Vinci Intuitive Surgical robotic system for antireflux surgery. In April 2003, we set up a pilot study to evaluate the efficacy of laparoscopic telerobotic surgery using the three-arm Da Vinci system. Optimal trocar positions, operating and setup times, conversion rate, intraoperative complications, and perioperative morbidity, as well as mortality rate, were analyzed. The median age was 53 years (range 25-74) in 118 patients (52 female/66 male). In 17 patients, an upside-down stomach- and in 101 GERD was surgical indication. The median operating time has been reduced from 105 min to 91 min after 40 procedures and setup time from 24.5 min to 10.4 min after 10 procedures. The system is safe and it seems to be superior to traditional laparoscopy during dissection in the esophageal hiatus region. This compensates long setup- and operating times. Disadvantages are the high costs, the time to master the setup/system and the necessity of exact trocar positioning.

  18. Comparison of individual and composite field analysis using array detector for Intensity Modulated Radiotherapy dose verification.

    PubMed

    Saminathan, Sathiyan; Chandraraj, Varatharaj; Sridhar, C H; Manickam, Ravikumar

    2012-01-01

    To compare the measured and calculated individual and composite field planar dose distribution of Intensity Modulated Radiotherapy plans. The measurements were performed in Clinac DHX linear accelerator with 6 MV photons using Matrixx device and a solid water phantom. The 20 brain tumor patients were selected for this study. The IMRT plan was carried out for all the patients using Eclipse treatment planning system. The verification plan was produced for every original plan using CT scan of Matrixx embedded in the phantom. Every verification field was measured by the Matrixx. The TPS calculated and measured dose distributions were compared for individual and composite fields. The percentage of gamma pixel match for the dose distribution patterns were evaluated using gamma histogram. The gamma pixel match was 95-98% for 41 fields (39%) and 98% for 59 fields (61%) with individual fields. The percentage of gamma pixel match was 95-98% for 5 patients and 98% for other 12 patients with composite fields. Three patients showed a gamma pixel match of less than 95%. The comparison of percentage gamma pixel match for individual and composite fields showed more than 2.5% variation for 6 patients, more than 1% variation for 4 patients, while the remaining 10 patients showed less than 1% variation. The individual and composite field measurements showed good agreement with TPS calculated dose distribution for the studied patients. The measurement and data analysis for individual fields is a time consuming process, the composite field analysis may be sufficient enough for smaller field dose distribution analysis with array detectors.

  19. The dosimetric impact of daily setup error on target volumes and surrounding normal tissue in the treatment of prostate cancer with intensity-modulated radiation therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Algan, Ozer, E-mail: oalgan@ouhsc.edu; Jamgade, Ambarish; Ali, Imad

    2012-01-01

    The purpose of this study was to evaluate the impact of daily setup error and interfraction organ motion on the overall dosimetric radiation treatment plans. Twelve patients undergoing definitive intensity-modulated radiation therapy (IMRT) treatments for prostate cancer were evaluated in this institutional review board-approved study. Each patient had fiducial markers placed into the prostate gland before treatment planning computed tomography scan. IMRT plans were generated using the Eclipse treatment planning system. Each patient was treated to a dose of 8100 cGy given in 45 fractions. In this study, we retrospectively created a plan for each treatment day that had amore » shift available. To calculate the dose, the patient would have received under this plan, we mathematically 'negated' the shift by moving the isocenter in the exact opposite direction of the shift. The individualized daily plans were combined to generate an overall plan sum. The dose distributions from these plans were compared with the treatment plans that were used to treat the patients. Three-hundred ninety daily shifts were negated and their corresponding plans evaluated. The mean isocenter shift based on the location of the fiducial markers was 3.3 {+-} 6.5 mm to the right, 1.6 {+-} 5.1 mm posteriorly, and 1.0 {+-} 5.0 mm along the caudal direction. The mean D95 doses for the prostate gland when setup error was corrected and uncorrected were 8228 and 7844 cGy (p < 0.002), respectively, and for the planning target volume (PTV8100) was 8089 and 7303 cGy (p < 0.001), respectively. The mean V95 values when patient setup was corrected and uncorrected were 99.9% and 87.3%, respectively, for the PTV8100 volume (p < 0.0001). At an individual patient level, the difference in the D95 value for the prostate volume could be >1200 cGy and for the PTV8100 could approach almost 2000 cGy when comparing corrected against uncorrected plans. There was no statistically significant difference in the D35 parameter for the surrounding normal tissue except for the dose received by the penile bulb and the right hip. Our dosimetric evaluation suggests significant underdosing with inaccurate target localization and emphasizes the importance of accurate patient setup and target localization. Further studies are needed to evaluate the impact of intrafraction organ motion, rotation, and deformation on doses delivered to target volumes.« less

  20. Development of Real Time Implementation of 5/5 Rule based Fuzzy Logic Controller Shunt Active Power Filter for Power Quality Improvement

    NASA Astrophysics Data System (ADS)

    Puhan, Pratap Sekhar; Ray, Pravat Kumar; Panda, Gayadhar

    2016-12-01

    This paper presents the effectiveness of 5/5 Fuzzy rule implementation in Fuzzy Logic Controller conjunction with indirect control technique to enhance the power quality in single phase system, An indirect current controller in conjunction with Fuzzy Logic Controller is applied to the proposed shunt active power filter to estimate the peak reference current and capacitor voltage. Current Controller based pulse width modulation (CCPWM) is used to generate the switching signals of voltage source inverter. Various simulation results are presented to verify the good behaviour of the Shunt active Power Filter (SAPF) with proposed two levels Hysteresis Current Controller (HCC). For verification of Shunt Active Power Filter in real time, the proposed control algorithm has been implemented in laboratory developed setup in dSPACE platform.

  1. Device-Independent Certification of a Nonprojective Qubit Measurement

    NASA Astrophysics Data System (ADS)

    Gómez, Esteban S.; Gómez, Santiago; González, Pablo; Cañas, Gustavo; Barra, Johanna F.; Delgado, Aldo; Xavier, Guilherme B.; Cabello, Adán; Kleinmann, Matthias; Vértesi, Tamás; Lima, Gustavo

    2016-12-01

    Quantum measurements on a two-level system can have more than two independent outcomes, and in this case, the measurement cannot be projective. Measurements of this general type are essential to an operational approach to quantum theory, but so far, the nonprojective character of a measurement can only be verified experimentally by already assuming a specific quantum model of parts of the experimental setup. Here, we overcome this restriction by using a device-independent approach. In an experiment on pairs of polarization-entangled photonic qubits we violate by more than 8 standard deviations a Bell-like correlation inequality that is valid for all sets of two-outcome measurements in any dimension. We combine this with a device-independent verification that the system is best described by two qubits, which therefore constitutes the first device-independent certification of a nonprojective quantum measurement.

  2. Coupled Facility-Payload Vibration Modeling Improvements

    NASA Technical Reports Server (NTRS)

    Carnahan, Timothy M.; Kaiser, Michael A.

    2015-01-01

    A major phase of aerospace hardware verification is vibration testing. The standard approach for such testing is to use a shaker to induce loads into the payload. In preparation for vibration testing at National Aeronautics and Space Administration/Goddard Space Flight Center an analysis is performed to assess the responses of the payload. A new method of modeling the test is presented that takes into account dynamic interactions between the facility and the payload. This dynamic interaction has affected testing in the past, but been ignored or adjusted for during testing. By modeling the combined dynamics of the facility and test article (payload) it is possible to improve the prediction of hardware responses. Many aerospace test facilities work in similar way to those at NASA/Goddard Space Flight Center. Lessons learned here should be applicable to other test facilities with similar setups.

  3. An algebraic iterative reconstruction technique for differential X-ray phase-contrast computed tomography.

    PubMed

    Fu, Jian; Schleede, Simone; Tan, Renbo; Chen, Liyuan; Bech, Martin; Achterhold, Klaus; Gifford, Martin; Loewen, Rod; Ruth, Ronald; Pfeiffer, Franz

    2013-09-01

    Iterative reconstruction has a wide spectrum of proven advantages in the field of conventional X-ray absorption-based computed tomography (CT). In this paper, we report on an algebraic iterative reconstruction technique for grating-based differential phase-contrast CT (DPC-CT). Due to the differential nature of DPC-CT projections, a differential operator and a smoothing operator are added to the iterative reconstruction, compared to the one commonly used for absorption-based CT data. This work comprises a numerical study of the algorithm and its experimental verification using a dataset measured at a two-grating interferometer setup. Since the algorithm is easy to implement and allows for the extension to various regularization possibilities, we expect a significant impact of the method for improving future medical and industrial DPC-CT applications. Copyright © 2012. Published by Elsevier GmbH.

  4. Annual verifications--a tick-box exercise?

    PubMed

    Walker, Gwen; Williams, David

    2014-09-01

    With the onus on healthcare providers and their staff to protect patients against all elements of 'avoidable harm' perhaps never greater, Gwen Walker, a highly experienced infection prevention control nurse specialist, and David Williams, MD of Approved Air, who has 30 years' experience in validation and verification of ventilation and ultraclean ventilation systems, examine changing requirements for, and trends in, operating theatre ventilation. Validation and verification reporting on such vital HVAC equipment should not, they argue, merely be viewed as a 'tick-box exercise'; it should instead 'comprehensively inform key stakeholders, and ultimately form part of clinical governance, thus protecting those ultimately named responsible for organisation-wide safety at Trust board level'.

  5. Local setup errors in image-guided radiotherapy for head and neck cancer patients immobilized with a custom-made device.

    PubMed

    Giske, Kristina; Stoiber, Eva M; Schwarz, Michael; Stoll, Armin; Muenter, Marc W; Timke, Carmen; Roeder, Falk; Debus, Juergen; Huber, Peter E; Thieke, Christian; Bendl, Rolf

    2011-06-01

    To evaluate the local positioning uncertainties during fractionated radiotherapy of head-and-neck cancer patients immobilized using a custom-made fixation device and discuss the effect of possible patient correction strategies for these uncertainties. A total of 45 head-and-neck patients underwent regular control computed tomography scanning using an in-room computed tomography scanner. The local and global positioning variations of all patients were evaluated by applying a rigid registration algorithm. One bounding box around the complete target volume and nine local registration boxes containing relevant anatomic structures were introduced. The resulting uncertainties for a stereotactic setup and the deformations referenced to one anatomic local registration box were determined. Local deformations of the patients immobilized using our custom-made device were compared with previously published results. Several patient positioning correction strategies were simulated, and the residual local uncertainties were calculated. The patient anatomy in the stereotactic setup showed local systematic positioning deviations of 1-4 mm. The deformations referenced to a particular anatomic local registration box were similar to the reported deformations assessed from patients immobilized with commercially available Aquaplast masks. A global correction, including the rotational error compensation, decreased the remaining local translational errors. Depending on the chosen patient positioning strategy, the remaining local uncertainties varied considerably. Local deformations in head-and-neck patients occur even if an elaborate, custom-made patient fixation method is used. A rotational error correction decreased the required margins considerably. None of the considered correction strategies achieved perfect alignment. Therefore, weighting of anatomic subregions to obtain the optimal correction vector should be investigated in the future. Copyright © 2011 Elsevier Inc. All rights reserved.

  6. Three independent one-dimensional margins for single-fraction frameless stereotactic radiosurgery brain cases using CBCT

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Zhang, Qinghui; Chan, Maria F.; Burman, Chandra

    2013-12-15

    Purpose: Setting a proper margin is crucial for not only delivering the required radiation dose to a target volume, but also reducing the unnecessary radiation to the adjacent organs at risk. This study investigated the independent one-dimensional symmetric and asymmetric margins between the clinical target volume (CTV) and the planning target volume (PTV) for linac-based single-fraction frameless stereotactic radiosurgery (SRS).Methods: The authors assumed a Dirac delta function for the systematic error of a specific machine and a Gaussian function for the residual setup errors. Margin formulas were then derived in details to arrive at a suitable CTV-to-PTV margin for single-fractionmore » frameless SRS. Such a margin ensured that the CTV would receive the prescribed dose in 95% of the patients. To validate our margin formalism, the authors retrospectively analyzed nine patients who were previously treated with noncoplanar conformal beams. Cone-beam computed tomography (CBCT) was used in the patient setup. The isocenter shifts between the CBCT and linac were measured for a Varian Trilogy linear accelerator for three months. For each plan, the authors shifted the isocenter of the plan in each direction by ±3 mm simultaneously to simulate the worst setup scenario. Subsequently, the asymptotic behavior of the CTV V{sub 80%} for each patient was studied as the setup error approached the CTV-PTV margin.Results: The authors found that the proper margin for single-fraction frameless SRS cases with brain cancer was about 3 mm for the machine investigated in this study. The isocenter shifts between the CBCT and the linac remained almost constant over a period of three months for this specific machine. This confirmed our assumption that the machine systematic error distribution could be approximated as a delta function. This definition is especially relevant to a single-fraction treatment. The prescribed dose coverage for all the patients investigated was 96.1%± 5.5% with an extreme 3-mm setup error in all three directions simultaneously. It was found that the effect of the setup error on dose coverage was tumor location dependent. It mostly affected the tumors located in the posterior part of the brain, resulting in a minimum coverage of approximately 72%. This was entirely due to the unique geometry of the posterior head.Conclusions: Margin expansion formulas were derived for single-fraction frameless SRS such that the CTV would receive the prescribed dose in 95% of the patients treated for brain cancer. The margins defined in this study are machine-specific and account for nonzero mean systematic error. The margin for single-fraction SRS for a group of machines was also derived in this paper.« less

  7. The Iowa new practice model: Advancing technician roles to increase pharmacists' time to provide patient care services.

    PubMed

    Andreski, Michael; Myers, Megan; Gainer, Kate; Pudlo, Anthony

    Determine the effects of an 18-month pilot project using tech-check-tech in 7 community pharmacies on 1) rate of dispensing errors not identified during refill prescription final product verification; 2) pharmacist workday task composition; and 3) amount of patient care services provided and the reimbursement status of those services. Pretest-posttest quasi-experimental study where baseline and study periods were compared. Pharmacists and pharmacy technicians in 7 community pharmacies in Iowa. The outcome measures were 1) percentage of technician verified refill prescriptions where dispensing errors were not identified on final product verification; 2) percentage of time spent by pharmacists in dispensing, management, patient care, practice development, and other activities; 3) the number of pharmacist patient care services provided per pharmacist hours worked; and 4) percentage of time that technician product verification was used. There was no significant difference in overall errors (0.2729% vs. 0.5124%, P = 0.513), patient safety errors (0.0525% vs. 0.0651%, P = 0.837), or administrative errors (0.2204% vs. 0.4784%, P = 0.411). Pharmacist's time in dispensing significantly decreased (67.3% vs. 49.06%, P = 0.005), and time in direct patient care (19.96% vs. 34.72%, P = 0.003), increased significantly. Time in other activities did not significantly change. Reimbursable services per pharmacist hour (0.11 vs. 0.30, P = 0.129), did not significantly change. Non-reimbursable services increased significantly (2.77 vs. 4.80, P = 0.042). Total services significantly increased (2.88 vs. 5.16, P = 0.044). Pharmacy technician product verification of refill prescriptions preserved dispensing safety while significantly increasing the time spent in delivery of pharmacist provided patient care services. The total number of pharmacist services provided per hour also increased significantly, driven primarily by a significant increase in the number of non-reimbursed services. This was mostly likely due to the increased time available to provide patient care. Reimbursed services per hour did not increase significantly mostly likely due to lack of payers. Copyright © 2018 American Pharmacists Association®. Published by Elsevier Inc. All rights reserved.

  8. SU-F-T-647: Linac-Based Stereotactic Radiosurgery (SRS) in the Treatment of Trigeminal Neuralgia: Detailed Description of SRS Procedural Technique and Reported Clinical Outcomes

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Pokhrel, D; Sood, S; Badkul, R

    Purpose: SRS is an effective non-invasive alternative treatment modality with minimal-toxicity used to treat patients with medically/surgically refractory trigeminal neuralgia root(TNR) or those who may not tolerate surgical intervention. We present our linac-based SRS procedure for TNR treatment and simultaneously report our clinical outcomes. Methods: Twenty-eight TNR-patients treated with frame-based SRS at our institution (2009–2015) with a single-fraction point-dose of 60-80Gy to TNR were included in this IRB-approved study. Experienced neurosurgeon and radiation oncologist delineated the TNR on 1.0mm thin 3D-FIESTA-MRI that was co-registered with 0.7mm thin planning-CT. Treatment plans were generated in iPlan (BrainLAB) with a 4-mm diameter conemore » using 79 arcs with differential-weighting for Novalis-TX 6MV-SRS(1000MU/min) beam and optimized to minimize brainstem dose. Winston-Lutz test was performed before each treatment delivery with sub-millimeter isocenter accuracy. Quality assurance of frame placement was maintained by helmet-bobble-measurement before simulation-CT and before patient setup at treatment couch. OBI-CBCT scan was performed for patient setup verification without applying shifts. On clinical follow up, treatment response was assessed using Barrow Neurological Institute Pain Intensity Score(BNI-score:I–V). Results: 26/28 TNR-patients (16-males/10-females) who were treated with following single-fraction point-dose to isocenter: 80Gy(n=22),75Gy(n=1),70Gy(n=2) and 60Gy(n=1, re-treatment) were followed up. Median follow-up interval was 8.5-months (ranged:1–48.5months). Median age was 70-yr (ranged:43–93-yr). Right/left TNR ratio was 15/11. Delivered total # of average MUs was 19034±1204. Average beam-on-time: 19.0±1.3min. Brainstem max-dose and dose to 0.5cc were 13.3±2.4Gy (ranged:8.1–16.5Gy) and 3.6±0.4Gy (ranged:3.0–4.9Gy). On average, max-dose to optic-apparatus was ≤1.2Gy. Mean value of max-dose to eyes/lens was 0.26Gy/0.11Gy. Overall, 20-patients (77%) responded to treatment: 5(19%) achieved complete pain relief without medication (BNI score: I); 5(19%) had no-pain, decreased medication (BNI-score:II); 2(7.7%) had no-pain, but, continued medication (BNI-score:IIIA), and 8(30.8%) had pain that was well controlled by medication (BNI-score: IIIB). Six-patients (23.0%) did not respond to treatment (BNI-score:IV–V). Neither cranial nerve deficit nor radio-necrosis of temporal lobe was clinically observed. Conclusion: Linac-based SRS for medically/surgically refractory TNR provided an effective treatment option for pain resolution/control with very minimal if any normal tissue toxicity. Longer follow up of these patients is anticipated/needed to confirm our observations.« less

  9. The Maximal Oxygen Uptake Verification Phase: a Light at the End of the Tunnel?

    PubMed

    Schaun, Gustavo Z

    2017-12-08

    Commonly performed during an incremental test to exhaustion, maximal oxygen uptake (V̇O 2max ) assessment has become a recurring practice in clinical and experimental settings. To validate the test, several criteria were proposed. In this context, the plateau in oxygen uptake (V̇O 2 ) is inconsistent in its frequency, reducing its usefulness as a robust method to determine "true" V̇O 2max . Moreover, secondary criteria previously suggested, such as expiratory exchange ratios or percentages of maximal heart rate, are highly dependent on protocol design and often are achieved at V̇O 2 percentages well below V̇O 2max . Thus, an alternative method termed verification phase was proposed. Currently, it is clear that the verification phase can be a practical and sensitive method to confirm V̇O 2max ; however, procedures to conduct it are not standardized across the literature and no previous research tried to summarize how it has been employed. Therefore, in this review the knowledge on the verification phase was updated, while suggestions on how it can be performed (e.g. intensity, duration, recovery) were provided according to population and protocol design. Future studies should focus to identify a verification protocol feasible for different populations and to compare square-wave and multistage verification phases. Additionally, studies assessing verification phases in different patient populations are still warranted.

  10. Evaluation of the accuracy and clinical practicality of a calculation system for patient positional displacement in carbon ion radiotherapy at five sites.

    PubMed

    Kubota, Yoshiki; Hayashi, Hayato; Abe, Satoshi; Souda, Saki; Okada, Ryosuke; Ishii, Takayoshi; Tashiro, Mutsumi; Torikoshi, Masami; Kanai, Tatsuaki; Ohno, Tatsuya; Nakano, Takashi

    2018-03-01

    We developed a system for calculating patient positional displacement between digital radiography images (DRs) and digitally reconstructed radiography images (DRRs) to reduce patient radiation exposure, minimize individual differences between radiological technologists in patient positioning, and decrease positioning time. The accuracy of this system at five sites was evaluated with clinical data from cancer patients. The dependence of calculation accuracy on the size of the region of interest (ROI) and initial position was evaluated for clinical use. For a preliminary verification, treatment planning and positioning data from eight setup patterns using a head and neck phantom were evaluated. Following this, data from 50 patients with prostate, lung, head and neck, liver, or pancreatic cancer (n = 10 each) were evaluated. Root mean square errors (RMSEs) between the results calculated by our system and the reference positions were assessed. The reference positions were manually determined by two radiological technologists to best-matching positions with orthogonal DRs and DRRs in six axial directions. The ROI size dependence was evaluated by comparing RMSEs for three different ROI sizes. Additionally, dependence on initial position parameters was evaluated by comparing RMSEs for four position patterns. For the phantom study, the average (± standard deviation) translation error was 0.17 ± 0.05, rotation error was 0.17 ± 0.07, and ΔD was 0.14 ± 0.05. Using the optimal ROI size for each patient site, all cases of prostate, lung, and head and neck cancer with initial position parameters of 10 mm or under were acceptable in our tolerance. However, only four liver cancer cases and three pancreatic cancer cases were acceptable, because of low-reproducibility regions in the ROIs. Our system has clinical practicality for prostate, lung, and head and neck cancer cases. Additionally, our findings suggest ROI size dependence in some cases. © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  11. Monitoring Dosimetric Impact of Weight Loss With Kilovoltage (KV) Cone Beam CT (CBCT) During Parotid-Sparing IMRT and Concurrent Chemotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ho, Kean Fatt, E-mail: hokeanfatt@hotmail.com; Marchant, Tom; Moore, Chris

    2012-03-01

    Purpose: Parotid-sparing head-and-neck intensity-modulated radiotherapy (IMRT) can reduce long-term xerostomia. However, patients frequently experience weight loss and tumor shrinkage during treatment. We evaluate the use of kilovoltage (kV) cone beam computed tomography (CBCT) for dose monitoring and examine if the dosimetric impact of such changes on the parotid and critical neural structures warrants replanning during treatment. Methods and materials: Ten patients with locally advanced oropharyngeal cancer were treated with contralateral parotid-sparing IMRT concurrently with platinum-based chemotherapy. Mean doses of 65 Gy and 54 Gy were delivered to clinical target volume (CTV)1 and CTV2, respectively, in 30 daily fractions. CBCT wasmore » prospectively acquired weekly. Each CBCT was coregistered with the planned isocenter. The spinal cord, brainstem, parotids, larynx, and oral cavity were outlined on each CBCT. Dose distributions were recalculated on the CBCT after correcting the gray scale to provide accurate Hounsfield calibration, using the original IMRT plan configuration. Results: Planned contralateral parotid mean doses were not significantly different to those delivered during treatment (p > 0.1). Ipsilateral and contralateral parotids showed a mean reduction in volume of 29.7% and 28.4%, respectively. There was no significant difference between planned and delivered maximum dose to the brainstem (p = 0.6) or spinal cord (p = 0.2), mean dose to larynx (p = 0.5) and oral cavity (p = 0.8). End-of-treatment mean weight loss was 7.5 kg (8.8% of baseline weight). Despite a {>=}10% weight loss in 5 patients, there was no significant dosimetric change affecting the contralateral parotid and neural structures. Conclusions: Although patient weight loss and parotid volume shrinkage was observed, overall, there was no significant excess dose to the organs at risk. No replanning was felt necessary for this patient cohort, but a larger patient sample will be investigated to further confirm these results. Nevertheless, kilovoltage CBCT is a valuable tool for patient setup verification and monitoring of dosimetric variation during radiotherapy.« less

  12. The National Center for Oncological Hadron Therapy: status of the project and future clinical use of the facility.

    PubMed

    Orecchia, Roberto; Fossati, Piero; Rossi, Sandro

    2009-01-01

    Hadron therapy is an advanced radiotherapy technique that employs charged particle beams. Several particles (pions, oxygen, neon and helium ions) have been investigated in the past, but at present only protons and carbon ions are used in clinical practice. Hadron therapy has been used for more than 50 years, more than 50,000 patients have been treated worldwide, and many new facilities are being built. Indications are still a matter of debate. The Italian National Center for Oncological Hadron Therapy (CNAO) is under construction in Pavia and will begin to treat patients in the near future. The CNAO will be a center capable of using both protons and carbon ions. In the first phase, three rooms with vertical and horizontal fixed beams will be available, subsequently the center will be upgraded with two more rooms equipped with a rotating gantry. The facility will use active scanning delivery systems and state-of-the-art immobilization and setup verification devices. One additional room will be devoted to physical and radiobiological research. The CNAO will be a high-patient-throughput facility capable of treating more than 3,000 patients per year. Seven areas of interest have been identified: lung cancer, liver cancer, head and neck malignancies, pediatric solid cancers, eye tumors, sarcoma and central nervous system cancers. A disease-specific working group has been created for each area and has defined selection criteria and protocols to be used at the CNAO. Two more working groups are being set up on gynecological and digestive (pancreas, biliary tract and rectum) tumors. All the patients will participate in clinical trials to establish with sound evidence the real indications for hadron therapy. National and international cooperation networks are being set up to facilitate patient referral and follow-up. A medical service is already operative to assist patients and in selected case to refer them abroad. The CNAO will be the only carbon ion facility in Italy and will have an international basin. Close cooperation with existing oncological centers is of paramount importance to fully exploit its potential.

  13. Experimental verification of stopping-power prediction from single- and dual-energy computed tomography in biological tissues

    NASA Astrophysics Data System (ADS)

    Möhler, Christian; Russ, Tom; Wohlfahrt, Patrick; Elter, Alina; Runz, Armin; Richter, Christian; Greilich, Steffen

    2018-01-01

    An experimental setup for consecutive measurement of ion and x-ray absorption in tissue or other materials is introduced. With this setup using a 3D-printed sample container, the reference stopping-power ratio (SPR) of materials can be measured with an uncertainty of below 0.1%. A total of 65 porcine and bovine tissue samples were prepared for measurement, comprising five samples each of 13 tissue types representing about 80% of the total body mass (three different muscle and fatty tissues, liver, kidney, brain, heart, blood, lung and bone). Using a standard stoichiometric calibration for single-energy CT (SECT) as well as a state-of-the-art dual-energy CT (DECT) approach, SPR was predicted for all tissues and then compared to the measured reference. With the SECT approach, the SPRs of all tissues were predicted with a mean error of (-0.84  ±  0.12)% and a mean absolute error of (1.27  ±  0.12)%. In contrast, the DECT-based SPR predictions were overall consistent with the measured reference with a mean error of (-0.02  ±  0.15)% and a mean absolute error of (0.10  ±  0.15)%. Thus, in this study, the potential of DECT to decrease range uncertainty could be confirmed in biological tissue.

  14. Intrafraction Motion in Stereotactic Body Radiation Therapy for Non-Small Cell Lung Cancer: Intensity Modulated Radiation Therapy Versus Volumetric Modulated Arc Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rossi, Maddalena M.G.; Peulen, Heike M.U.; Belderbos, Josè S.A.

    Purpose: Stereotactic body radiation therapy (SBRT) for early-stage inoperable non-small cell lung cancer (NSCLC) patients delivers high doses that require high-precision treatment. Typically, image guidance is used to minimize day-to-day target displacement, but intrafraction position variability is often not corrected. Currently, volumetric modulated arc therapy (VMAT) is replacing intensity modulated radiation therapy (IMRT) in many departments because of its shorter delivery time. This study aimed to evaluate whether intrafraction variation in VMAT patients is reduced in comparison with patients treated with IMRT. Methods and Materials: NSCLC patients (197 IMRT and 112 VMAT) treated with a frameless SBRT technique to amore » prescribed dose of 3 × 18 Gy were evaluated. Image guidance for both techniques was identical: pretreatment cone beam computed tomography (CBCT) (CBCT{sub precorr}) for setup correction followed immediately before treatment by postcorrection CBCT (CBCT{sub postcorr}) for verification. Then, after either a noncoplanar IMRT technique or a VMAT technique, a posttreatment (CBCT{sub postRT}) scan was acquired. The CBCT{sub postRT} and CBCT{sub postcorr} scans were then used to evaluate intrafraction motion. Treatment delivery times, systematic (Σ) and random (σ) intrafraction variations, and associated planning target volume (PTV) margins were calculated. Results: The median treatment delivery time was significantly reduced by 20 minutes (range, 32-12 minutes) using VMAT compared with noncoplanar IMRT. Intrafraction tumor motion was significantly larger for IMRT in all directions up to 0.5 mm systematic (Σ) and 0.7 mm random (σ). The required PTV margins for IMRT and VMAT differed by less than 0.3 mm. Conclusion: VMAT-based SBRT for NSCLC was associated with significantly shorter delivery times and correspondingly smaller intrafraction motion compared with noncoplanar IMRT. However, the impact on the required PTV margin was small.« less

  15. Dosimetric verification for intensity-modulated arc therapy plans by use of 2D diode array, radiochromic film and radiosensitive polymer gel.

    PubMed

    Hayashi, Naoki; Malmin, Ryan L; Watanabe, Yoichi

    2014-05-01

    Several tools are used for the dosimetric verification of intensity-modulated arc therapy (IMAT) treatment delivery. However, limited information is available for composite on-line evaluation of these tools. The purpose of this study was to evaluate the dosimetric verification of IMAT treatment plans using a 2D diode array detector (2D array), radiochromic film (RCF) and radiosensitive polymer gel dosimeter (RPGD). The specific verification plans were created for IMAT for two prostate cancer patients by use of the clinical treatment plans. Accordingly, the IMAT deliveries were performed with the 2D array on a gantry-mounting device, RCF in a cylindrical acrylic phantom, and the RPGD in two cylindrical phantoms. After the irradiation, the planar dose distributions from the 2D array and the RCFs, and the 3D dose distributions from the RPGD measurements were compared with the calculated dose distributions using the gamma analysis method (3% dose difference and 3-mm distance-to-agreement criterion), dose-dependent dose difference diagrams, dose difference histograms, and isodose distributions. The gamma passing rates of 2D array, RCFs and RPGD for one patient were 99.5%, 96.5% and 93.7%, respectively; the corresponding values for the second patient were 97.5%, 92.6% and 92.9%. Mean percentage differences between the RPGD measured and calculated doses in 3D volumes containing PTVs were -0.29 ± 7.1% and 0.97 ± 7.6% for the two patients, respectively. In conclusion, IMAT prostate plans can be delivered with high accuracy, although the 3D measurements indicated less satisfactory agreement with the treatment plans, mainly due to the dosimetric inaccuracy in low-dose regions of the RPGD measurements.

  16. Imperceptible watermarking for security of fundus images in tele-ophthalmology applications and computer-aided diagnosis of retina diseases.

    PubMed

    Singh, Anushikha; Dutta, Malay Kishore

    2017-12-01

    The authentication and integrity verification of medical images is a critical and growing issue for patients in e-health services. Accurate identification of medical images and patient verification is an essential requirement to prevent error in medical diagnosis. The proposed work presents an imperceptible watermarking system to address the security issue of medical fundus images for tele-ophthalmology applications and computer aided automated diagnosis of retinal diseases. In the proposed work, patient identity is embedded in fundus image in singular value decomposition domain with adaptive quantization parameter to maintain perceptual transparency for variety of fundus images like healthy fundus or disease affected image. In the proposed method insertion of watermark in fundus image does not affect the automatic image processing diagnosis of retinal objects & pathologies which ensure uncompromised computer-based diagnosis associated with fundus image. Patient ID is correctly recovered from watermarked fundus image for integrity verification of fundus image at the diagnosis centre. The proposed watermarking system is tested in a comprehensive database of fundus images and results are convincing. results indicate that proposed watermarking method is imperceptible and it does not affect computer vision based automated diagnosis of retinal diseases. Correct recovery of patient ID from watermarked fundus image makes the proposed watermarking system applicable for authentication of fundus images for computer aided diagnosis and Tele-ophthalmology applications. Copyright © 2017 Elsevier B.V. All rights reserved.

  17. Limited Impact of Setup and Range Uncertainties, Breathing Motion, and Interplay Effects in Robustly Optimized Intensity Modulated Proton Therapy for Stage III Non-small Cell Lung Cancer.

    PubMed

    Inoue, Tatsuya; Widder, Joachim; van Dijk, Lisanne V; Takegawa, Hideki; Koizumi, Masahiko; Takashina, Masaaki; Usui, Keisuke; Kurokawa, Chie; Sugimoto, Satoru; Saito, Anneyuko I; Sasai, Keisuke; Van't Veld, Aart A; Langendijk, Johannes A; Korevaar, Erik W

    2016-11-01

    To investigate the impact of setup and range uncertainties, breathing motion, and interplay effects using scanning pencil beams in robustly optimized intensity modulated proton therapy (IMPT) for stage III non-small cell lung cancer (NSCLC). Three-field IMPT plans were created using a minimax robust optimization technique for 10 NSCLC patients. The plans accounted for 5- or 7-mm setup errors with ±3% range uncertainties. The robustness of the IMPT nominal plans was evaluated considering (1) isotropic 5-mm setup errors with ±3% range uncertainties; (2) breathing motion; (3) interplay effects; and (4) a combination of items 1 and 2. The plans were calculated using 4-dimensional and average intensity projection computed tomography images. The target coverage (TC, volume receiving 95% of prescribed dose) and homogeneity index (D2 - D98, where D2 and D98 are the least doses received by 2% and 98% of the volume) for the internal clinical target volume, and dose indexes for lung, esophagus, heart and spinal cord were compared with that of clinical volumetric modulated arc therapy plans. The TC and homogeneity index for all plans were within clinical limits when considering the breathing motion and interplay effects independently. The setup and range uncertainties had a larger effect when considering their combined effect. The TC decreased to <98% (clinical threshold) in 3 of 10 patients for robust 5-mm evaluations. However, the TC remained >98% for robust 7-mm evaluations for all patients. The organ at risk dose parameters did not significantly vary between the respective robust 5-mm and robust 7-mm evaluations for the 4 error types. Compared with the volumetric modulated arc therapy plans, the IMPT plans showed better target homogeneity and mean lung and heart dose parameters reduced by about 40% and 60%, respectively. In robustly optimized IMPT for stage III NSCLC, the setup and range uncertainties, breathing motion, and interplay effects have limited impact on target coverage, dose homogeneity, and organ-at-risk dose parameters. Copyright © 2016 Elsevier Inc. All rights reserved.

  18. SU-F-J-21: Clinical Evaluation of Surface Scanning Systems in Different Treatment Locations

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Moser, T; Karger, C; Stefanowicz, S

    Purpose: To reduce imaging dose in fractionated IGRT, the ability of optical surface imaging systems (OSIS) to detect setup errors was tested. Therefore, clinical studies to evaluate for different treatment locations setup corrections derived by OSIS in comparison with x-ray image guidance in fractionated radiation therapy was performed. Methods: The setup correction accuracy of an OSIS system (AlignRT, VisionRT, London, UK) will be analysed for the 4 tumour locations Pelvis, Upper Abdomen, Thorax and Breast, 20 patients for each location in comparison to a different system (Sentinel, C-RAD, SE). For each patient, the setup corrections of the cone-beam computed tomographymore » (CBCT) of an Elekta Versa HD linear accelerator (Elekta, Crawley, UK) is considered as gold-standard and then compared with those of the OSIS for the first ten fractions retrospectively. There were no clinical decisions made based on the surrogate system. For the OSIS, the reference surface is highly important as it represents the actual ground truth. It can be obtained either with the system itself or the surface structure delineated in the planning CT can be imported via DICOM interface. In this paper, the first results for the treatment region thorax are presented. The reference image modalities were compared. Results: Table 1 displays the difference between the setup corrections obtained with OSIS and CBCT in lateral (LAT), longitudinal (LNG) and vertical (VRT) direction for the DICOM reference image. While the median deviations are within a few millimeters, some outliers showed large deviations. Generally, the mean deviation as well as the spread was smallest in lateral and largest in vertical direction. Conclusion: Although the system allows fast, simple and non-invasive determination of setup corrections, it should be evaluated treatment region dependant. Therefore, the study is ongoing. The application of OSIS may help to reduce the imaging dose for the patient. We gratefully acknowledge the support by our colleagues from the Radiological University Clinic Heidelberg, where the study was performed. This work was funded by the Federal Ministry of Education and Research (BMBF) Germany, grant number 01IB13001B.« less

  19. 4D laser camera for accurate patient positioning, collision avoidance, image fusion and adaptive approaches during diagnostic and therapeutic procedures.

    PubMed

    Brahme, Anders; Nyman, Peter; Skatt, Björn

    2008-05-01

    A four-dimensional (4D) laser camera (LC) has been developed for accurate patient imaging in diagnostic and therapeutic radiology. A complementary metal-oxide semiconductor camera images the intersection of a scanned fan shaped laser beam with the surface of the patient and allows real time recording of movements in a three-dimensional (3D) or four-dimensional (4D) format (3D +time). The LC system was first designed as an accurate patient setup tool during diagnostic and therapeutic applications but was found to be of much wider applicability as a general 4D photon "tag" for the surface of the patient in different clinical procedures. It is presently used as a 3D or 4D optical benchmark or tag for accurate delineation of the patient surface as demonstrated for patient auto setup, breathing and heart motion detection. Furthermore, its future potential applications in gating, adaptive therapy, 3D or 4D image fusion between most imaging modalities and image processing are discussed. It is shown that the LC system has a geometrical resolution of about 0, 1 mm and that the rigid body repositioning accuracy is about 0, 5 mm below 20 mm displacements, 1 mm below 40 mm and better than 2 mm at 70 mm. This indicates a slight need for repeated repositioning when the initial error is larger than about 50 mm. The positioning accuracy with standard patient setup procedures for prostate cancer at Karolinska was found to be about 5-6 mm when independently measured using the LC system. The system was found valuable for positron emission tomography-computed tomography (PET-CT) in vivo tumor and dose delivery imaging where it potentially may allow effective correction for breathing artifacts in 4D PET-CT and image fusion with lymph node atlases for accurate target volume definition in oncology. With a LC system in all imaging and radiation therapy rooms, auto setup during repeated diagnostic and therapeutic procedures may save around 5 min per session, increase accuracy and allow efficient image fusion between all imaging modalities employed.

  20. MRM for the verification of cancer biomarker proteins: recent applications to human plasma and serum.

    PubMed

    Chambers, Andrew G; Percy, Andrew J; Simon, Romain; Borchers, Christoph H

    2014-04-01

    Accurate cancer biomarkers are needed for early detection, disease classification, prediction of therapeutic response and monitoring treatment. While there appears to be no shortage of candidate biomarker proteins, a major bottleneck in the biomarker pipeline continues to be their verification by enzyme linked immunosorbent assays. Multiple reaction monitoring (MRM), also known as selected reaction monitoring, is a targeted mass spectrometry approach to protein quantitation and is emerging to bridge the gap between biomarker discovery and clinical validation. Highly multiplexed MRM assays are readily configured and enable simultaneous verification of large numbers of candidates facilitating the development of biomarker panels which can increase specificity. This review focuses on recent applications of MRM to the analysis of plasma and serum from cancer patients for biomarker verification. The current status of this approach is discussed along with future directions for targeted mass spectrometry in clinical biomarker validation.

  1. Self-verification and social anxiety: preference for negative social feedback and low social self-esteem.

    PubMed

    Valentiner, David P; Skowronski, John J; McGrath, Patrick B; Smith, Sarah A; Renner, Kerry A

    2011-10-01

    A self-verification model of social anxiety views negative social self-esteem as a core feature of social anxiety. This core feature is proposed to be maintained through self-verification processes, such as by leading individuals with negative social self-esteem to prefer negative social feedback. This model is tested in two studies. In Study 1, questionnaires were administered to a college sample (N = 317). In Study 2, questionnaires were administered to anxiety disordered patients (N = 62) before and after treatment. Study 1 developed measures of preference for negative social feedback and social self-esteem, and provided evidence of their incremental validity in a college sample. Study 2 found that these two variables are not strongly related to fears of evaluation, are relatively unaffected by a treatment that targets such fears, and predict residual social anxiety following treatment. Overall, these studies provide preliminary evidence for a self-verification model of social anxiety.

  2. Verification of monitor unit calculations for non-IMRT clinical radiotherapy: report of AAPM Task Group 114.

    PubMed

    Stern, Robin L; Heaton, Robert; Fraser, Martin W; Goddu, S Murty; Kirby, Thomas H; Lam, Kwok Leung; Molineu, Andrea; Zhu, Timothy C

    2011-01-01

    The requirement of an independent verification of the monitor units (MU) or time calculated to deliver the prescribed dose to a patient has been a mainstay of radiation oncology quality assurance. The need for and value of such a verification was obvious when calculations were performed by hand using look-up tables, and the verification was achieved by a second person independently repeating the calculation. However, in a modern clinic using CT/MR/PET simulation, computerized 3D treatment planning, heterogeneity corrections, and complex calculation algorithms such as convolution/superposition and Monte Carlo, the purpose of and methodology for the MU verification have come into question. In addition, since the verification is often performed using a simpler geometrical model and calculation algorithm than the primary calculation, exact or almost exact agreement between the two can no longer be expected. Guidelines are needed to help the physicist set clinically reasonable action levels for agreement. This report addresses the following charges of the task group: (1) To re-evaluate the purpose and methods of the "independent second check" for monitor unit calculations for non-IMRT radiation treatment in light of the complexities of modern-day treatment planning. (2) To present recommendations on how to perform verification of monitor unit calculations in a modern clinic. (3) To provide recommendations on establishing action levels for agreement between primary calculations and verification, and to provide guidance in addressing discrepancies outside the action levels. These recommendations are to be used as guidelines only and shall not be interpreted as requirements.

  3. Polysomnography versus limited respiratory monitoring and nurse-led titration to optimise non-invasive ventilation set-up: a pilot randomised clinical trial.

    PubMed

    Patout, Maxime; Arbane, Gill; Cuvelier, Antoine; Muir, Jean Francois; Hart, Nicholas; Murphy, Patrick Brian

    2018-03-30

    Polysomnography (PSG) is recommended for non-invasive ventilation (NIV) set-up in patients with chronic respiratory failure. In this pilot randomised clinical trial, we compared the physiological effectiveness of NIV set-up guided by PSG to limited respiratory monitoring (LRM) and nurse-led titration in patients with COPD-obstructive sleep apnoea (OSA) overlap. The principal outcome of interest was change in daytime arterial partial pressure of carbon dioxide (PaCO 2 ) at 3 months. Fourteen patients with daytime PaCO 2 >6 kPa and body mass index >30 kg/m 2 were recruited. At 3 months, PaCO 2 was reduced by -0.88 kPa (95% CI -1.52 to -0.24 kPa) in the LRM group and by -0.36 kPa (95% CI -0.96 to 0.24 kPa) in the PSG group. These pilot data provide support to undertake a clinical trial investigating the clinical effectiveness of attended limited respiratory monitoring and PSG to establish NIV in patients with COPD-OSA overlap. Results, NCT02444806. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2018. All rights reserved. No commercial use is permitted unless otherwise expressly granted.

  4. An electromechanical, patient positioning system for head and neck radiotherapy

    NASA Astrophysics Data System (ADS)

    Ostyn, Mark; Dwyer, Thomas; Miller, Matthew; King, Paden; Sacks, Rachel; Cruikshank, Ross; Rosario, Melvin; Martinez, Daniel; Kim, Siyong; Yeo, Woon-Hong

    2017-09-01

    In cancer treatment with radiation, accurate patient setup is critical for proper dose delivery. Improper arrangement can lead to disease recurrence, permanent organ damage, or lack of disease control. While current immobilization equipment often helps for patient positioning, manual adjustment is required, involving iterative, time-consuming steps. Here, we present an electromechanical robotic system for improving patient setup in radiotherapy, specifically targeting head and neck cancer. This positioning system offers six degrees of freedom for a variety of applications in radiation oncology. An analytical calculation of inverse kinematics serves as fundamental criteria to design the system. Computational mechanical modeling and experimental study of radiotherapy compatibility and x-ray-based imaging demonstrates the device feasibility and reliability to be used in radiotherapy. An absolute positioning accuracy test in a clinical treatment room supports the clinical feasibility of the system.

  5. Dose verification with different ion chambers for SRT/SBRT plans

    NASA Astrophysics Data System (ADS)

    Durmus, I. F.; Tas, B.; Okumus, A.; Uzel, O. E.

    2017-02-01

    Verification of patient plan is very important in stereotactic treatments. VMAT plans were prepared with 6MV-FFF or 10MV-FFF energies for 25 intracranial and extracranial stereotactic patients. Absolute dose was measured for dose verification in each plans. Iba® CC01, Iba® CC04, Iba® CC13 ion chambers placed at a depth of 5cm in solid phantom (RW3). Also we scanned this phantom with ion chambers by Siemens® Biograph mCT. QA plans were prepared by transferring twenty five patient plans to phantom assemblies for three ion chambers. All plans were performed separately for three ion chambers at Elekta® Versa HD linear accelerator. Statistical analysis of results were made by Wilcoxon signed-rank test. Difference between dose values were determined %1.84±3.4 (p: 0.001) with Iba CC13 ion chamber, %1.80±3.4 (p: 0.002) with Iba CC04 ion chamber and %0.29±4.6 (p: 0.667) with Iba CC01 ion chamber. In stereotactic treatments, dosimetric uncertainty increases in small areas. We determined more accurate results with small sized detectors. Difference between TPS calculations and all measurements were founded lower than %2.

  6. Clinical verification in homeopathy and allergic conditions.

    PubMed

    Van Wassenhoven, Michel

    2013-01-01

    The literature on clinical research in allergic conditions treated with homeopathy includes a meta-analysis of randomised controlled trials (RCT) for hay fever with positive conclusions and two positive RCTs in asthma. Cohort surveys using validated Quality of Life questionnaires have shown improvement in asthma in children, general allergic conditions and skin diseases. Economic surveys have shown positive results in eczema, allergy, seasonal allergic rhinitis, asthma, food allergy and chronic allergic rhinitis. This paper reports clinical verification of homeopathic symptoms in all patients and especially in various allergic conditions in my own primary care practice. For preventive treatments in hay fever patients, Arsenicum album was the most effective homeopathic medicine followed by Nux vomica, Pulsatilla pratensis, Gelsemium, Sarsaparilla, Silicea and Natrum muriaticum. For asthma patients, Arsenicum iodatum appeared most effective, followed by Lachesis, Calcarea arsenicosa, Carbo vegetabilis and Silicea. For eczema and urticaria, Mezereum was most effective, followed by Lycopodium, Sepia, Arsenicum iodatum, Calcarea carbonica and Psorinum. The choice of homeopathic medicine depends on the presence of other associated symptoms and 'constitutional' features. Repertories should be updated by including results of such clinical verifications of homeopathic prescribing symptoms. Copyright © 2012 The Faculty of Homeopathy. Published by Elsevier Ltd. All rights reserved.

  7. Inverse probability weighting estimation of the volume under the ROC surface in the presence of verification bias.

    PubMed

    Zhang, Ying; Alonzo, Todd A

    2016-11-01

    In diagnostic medicine, the volume under the receiver operating characteristic (ROC) surface (VUS) is a commonly used index to quantify the ability of a continuous diagnostic test to discriminate between three disease states. In practice, verification of the true disease status may be performed only for a subset of subjects under study since the verification procedure is invasive, risky, or expensive. The selection for disease examination might depend on the results of the diagnostic test and other clinical characteristics of the patients, which in turn can cause bias in estimates of the VUS. This bias is referred to as verification bias. Existing verification bias correction in three-way ROC analysis focuses on ordinal tests. We propose verification bias-correction methods to construct ROC surface and estimate the VUS for a continuous diagnostic test, based on inverse probability weighting. By applying U-statistics theory, we develop asymptotic properties for the estimator. A Jackknife estimator of variance is also derived. Extensive simulation studies are performed to evaluate the performance of the new estimators in terms of bias correction and variance. The proposed methods are used to assess the ability of a biomarker to accurately identify stages of Alzheimer's disease. © 2016 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

  8. Runtime Verification of Pacemaker Functionality Using Hierarchical Fuzzy Colored Petri-nets.

    PubMed

    Majma, Negar; Babamir, Seyed Morteza; Monadjemi, Amirhassan

    2017-02-01

    Today, implanted medical devices are increasingly used for many patients and in case of diverse health problems. However, several runtime problems and errors are reported by the relevant organizations, even resulting in patient death. One of those devices is the pacemaker. The pacemaker is a device helping the patient to regulate the heartbeat by connecting to the cardiac vessels. This device is directed by its software, so any failure in this software causes a serious malfunction. Therefore, this study aims to a better way to monitor the device's software behavior to decrease the failure risk. Accordingly, we supervise the runtime function and status of the software. The software verification means examining limitations and needs of the system users by the system running software. In this paper, a method to verify the pacemaker software, based on the fuzzy function of the device, is presented. So, the function limitations of the device are identified and presented as fuzzy rules and then the device is verified based on the hierarchical Fuzzy Colored Petri-net (FCPN), which is formed considering the software limits. Regarding the experiences of using: 1) Fuzzy Petri-nets (FPN) to verify insulin pumps, 2) Colored Petri-nets (CPN) to verify the pacemaker and 3) To verify the pacemaker by a software agent with Petri-network based knowledge, which we gained during the previous studies, the runtime behavior of the pacemaker software is examined by HFCPN, in this paper. This is considered a developing step compared to the earlier work. HFCPN in this paper, compared to the FPN and CPN used in our previous studies reduces the complexity. By presenting the Petri-net (PN) in a hierarchical form, the verification runtime, decreased as 90.61% compared to the verification runtime in the earlier work. Since we need an inference engine in the runtime verification, we used the HFCPN to enhance the performance of the inference engine.

  9. MO-AB-BRA-03: Development of Novel Real Time in Vivo EPID Treatment Verification for Brachytherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fonseca, G; Podesta, M; Reniers, B

    2016-06-15

    Purpose: High Dose Rate (HDR) brachytherapy treatments are employed worldwide to treat a wide variety of cancers. However, in vivo dose verification remains a challenge with no commercial dosimetry system available to verify the treatment dose delivered to the patient. We propose a novel dosimetry system that couples an independent Monte Carlo (MC) simulation platform and an amorphous silicon Electronic Portal Imaging Device (EPID) to provide real time treatment verification. Methods: MC calculations predict the EPID response to the photon fluence emitted by the HDR source by simulating the patient, the source dwell positions and times, and treatment complexities suchmore » as tissue compositions/densities and different applicators. Simulated results are then compared against EPID measurements acquired with ∼0.14s time resolution which allows dose measurements for each dwell position. The EPID has been calibrated using an Ir-192 HDR source and experiments were performed using different phantoms, including tissue equivalent materials (PMMA, lung and bone). A source positioning accuracy of 0.2 mm, without including the afterloader uncertainty, was ensured using a robotic arm moving the source. Results: An EPID can acquire 3D Cartesian source positions and its response varies significantly due to differences in the material composition/density of the irradiated object, allowing detection of changes in patient geometry. The panel time resolution allows dose rate and dwell time measurements. Moreover, predicted EPID images obtained from clinical treatment plans provide anatomical information that can be related to the patient anatomy, mostly bone and air cavities, localizing the source inside of the patient using its anatomy as reference. Conclusion: Results obtained show the feasibility of the proposed dose verification system that is capable to verify all the brachytherapy treatment steps in real time providing data about treatment delivery quality and also applicator/structure motion during or between treatments.« less

  10. SU-E-T-602: Patient-Specific Online Dose Verification Based On Transmission Detector Measurements

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Thoelking, J; Yuvaraj, S; Jens, F

    Purpose: Intensity modulated radiotherapy requires a comprehensive quality assurance program in general and ideally independent verification of dose delivery. Since conventional 2D detector arrays allow only pre-treatment verification, there is a debate concerning the need of online dose verification. This study presents the clinical performance, including dosimetric plan verification in 2D as well as in 3D and the error detection abilities of a new transmission detector (TD) for online dose verification of 6MV photon beam. Methods: To validate the dosimetric performance of the new device, dose reconstruction based on TD measurements were compared to a conventional pre-treatment verification method (reference)more » and treatment planning system (TPS) for 18 IMRT and VMAT treatment plans. Furthermore, dose reconstruction inside the patient based on TD read-out was evaluated by comparing various dose volume indices and 3D gamma evaluations against independent dose computation and TPS. To investigate the sensitivity of the new device, different types of systematic and random errors for leaf positions and linac output were introduced in IMRT treatment sequences. Results: The 2D gamma index evaluation of transmission detector based dose reconstruction showed an excellent agreement for all IMRT and VMAT plans compared to reference measurements (99.3±1.2)% and TPS (99.1±0.7)%. Good agreement was also obtained for 3D dose reconstruction based on TD read-out compared to dose computation (mean gamma value of PTV = 0.27±0.04). Only a minimal dose underestimation within the target volume was observed when analyzing DVH indices (<1%). Positional errors in leaf banks larger than 1mm and errors in linac output larger than 2% could clearly identified with the TD. Conclusion: Since 2D and 3D evaluations for all IMRT and VMAT treatment plans were in excellent agreement with reference measurements and dose computation, the new TD is suitable to qualify for routine treatment plan verification. Funding Support, Disclosures, and Conflict of Interest: COIs: Frank Lohr: Elekta: research grant, travel grants, teaching honoraria IBA: research grant, travel grants, teaching honoraria, advisory board C-Rad: board honoraria, travel grants Frederik Wenz: Elekta: research grant, teaching honoraria, consultant, advisory board Zeiss: research grant, teaching honoraria, patent Hansjoerg Wertz: Elekta: research grant, teaching honoraria IBA: research grant.« less

  11. SU-F-J-192: A Quick and Effective Method to Validate Patient’s Daily Setup and Geometry Changes Prior to Proton Treatment Delivery Based On Water Equivalent Thickness Projection Imaging (WETPI) for Head Neck Cancer (HNC) Patient

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Liu, G; Qin, A; Zhang, J

    Purpose: With the implementation of Cone-beam Computed-Tomography (CBCT) in proton treatment, we introduces a quick and effective tool to verify the patient’s daily setup and geometry changes based on the Water-Equivalent-Thickness Projection-Image(WETPI) from individual beam angle. Methods: A bilateral head neck cancer(HNC) patient previously treated via VMAT was used in this study. The patient received 35 daily CBCT during the whole treatment and there is no significant weight change. The CT numbers of daily CBCTs were corrected by mapping the CT numbers from simulation CT via Deformable Image Registration(DIR). IMPT plan was generated using 4-field IMPT robust optimization (3.5% rangemore » and 3mm setup uncertainties) with beam angle 60, 135, 300, 225 degree. WETPI within CTV through all beam directions were calculated. 3%/3mm gamma index(GI) were used to provide a quantitative comparison between initial sim-CT and mapped daily CBCT. To simulate an extreme case where human error is involved, a couch bar was manually inserted in front of beam angle 225 degree of one CBCT. WETPI was compared in this scenario. Results: The average of GI passing rate of this patient from different beam angles throughout the treatment course is 91.5 ± 8.6. In the cases with low passing rate, it was found that the difference between shoulder and neck angle as well as the head rest often causes major deviation. This indicates that the most challenge in treating HNC is the setup around neck area. In the extreme case where a couch bar is accidently inserted in the beam line, GI passing rate drops to 52 from 95. Conclusion: WETPI and quantitative gamma analysis give clinicians, therapists and physicists a quick feedback of the patient’s setup accuracy or geometry changes. The tool could effectively avoid some human errors. Furthermore, this tool could be used potentially as an initial signal to trigger plan adaptation.« less

  12. SU-F-T-440: The Feasibility Research of Checking Cervical Cancer IMRT Pre- Treatment Dose Verification by Automated Treatment Planning Verification System

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Liu, X; Yin, Y; Lin, X

    Purpose: To assess the preliminary feasibility of automated treatment planning verification system in cervical cancer IMRT pre-treatment dose verification. Methods: The study selected randomly clinical IMRT treatment planning data for twenty patients with cervical cancer, all IMRT plans were divided into 7 fields to meet the dosimetric goals using a commercial treatment planning system(PianncleVersion 9.2and the EclipseVersion 13.5). The plans were exported to the Mobius 3D (M3D)server percentage differences of volume of a region of interest (ROI) and dose calculation of target region and organ at risk were evaluated, in order to validate the accuracy automated treatment planning verification system.more » Results: The difference of volume for Pinnacle to M3D was less than results for Eclipse to M3D in ROI, the biggest difference was 0.22± 0.69%, 3.5±1.89% for Pinnacle and Eclipse respectively. M3D showed slightly better agreement in dose of target and organ at risk compared with TPS. But after recalculating plans by M3D, dose difference for Pinnacle was less than Eclipse on average, results were within 3%. Conclusion: The method of utilizing the automated treatment planning system to validate the accuracy of plans is convenientbut the scope of differences still need more clinical patient cases to determine. At present, it should be used as a secondary check tool to improve safety in the clinical treatment planning.« less

  13. Visuo-acoustic stimulation that helps you to relax: A virtual reality setup for patients in the intensive care unit.

    PubMed

    Gerber, Stephan M; Jeitziner, Marie-Madlen; Wyss, Patric; Chesham, Alvin; Urwyler, Prabitha; Müri, René M; Jakob, Stephan M; Nef, Tobias

    2017-10-16

    After prolonged stay in an intensive care unit (ICU) patients often complain about cognitive impairments that affect health-related quality of life after discharge. The aim of this proof-of-concept study was to test the feasibility and effects of controlled visual and acoustic stimulation in a virtual reality (VR) setup in the ICU. The VR setup consisted of a head-mounted display in combination with an eye tracker and sensors to assess vital signs. The stimulation consisted of videos featuring natural scenes and was tested in 37 healthy participants in the ICU. The VR stimulation led to a reduction of heart rate (p = 0. 049) and blood pressure (p = 0.044). Fixation/saccade ratio (p < 0.001) was increased when a visual target was presented superimposed on the videos (reduced search activity), reflecting enhanced visual processing. Overall, the VR stimulation had a relaxing effect as shown in vital markers of physical stress and participants explored less when attending the target. Our study indicates that VR stimulation in ICU settings is feasible and beneficial for critically ill patients.

  14. Verification bias an underrecognized source of error in assessing the efficacy of medical imaging.

    PubMed

    Petscavage, Jonelle M; Richardson, Michael L; Carr, Robert B

    2011-03-01

    Diagnostic tests are validated by comparison against a "gold standard" reference test. When the reference test is invasive or expensive, it may not be applied to all patients. This can result in biased estimates of the sensitivity and specificity of the diagnostic test. This type of bias is called "verification bias," and is a common problem in imaging research. The purpose of our study is to estimate the prevalence of verification bias in the recent radiology literature. All issues of the American Journal of Roentgenology (AJR), Academic Radiology, Radiology, and European Journal of Radiology (EJR) between November 2006 and October 2009 were reviewed for original research articles mentioning sensitivity or specificity as endpoints. Articles were read to determine whether verification bias was present and searched for author recognition of verification bias in the design. During 3 years, these journals published 2969 original research articles. A total of 776 articles used sensitivity or specificity as an outcome. Of these, 211 articles demonstrated potential verification bias. The fraction of articles with potential bias was respectively 36.4%, 23.4%, 29.5%, and 13.4% for AJR, Academic Radiology, Radiology, and EJR. The total fraction of papers with potential bias in which the authors acknowledged this bias was 17.1%. Verification bias is a common and frequently unacknowledged source of error in efficacy studies of diagnostic imaging. Bias can often be eliminated by proper study design. When it cannot be eliminated, it should be estimated and acknowledged. Published by Elsevier Inc.

  15. Color structured light system of chest wall motion measurement for respiratory volume evaluation

    NASA Astrophysics Data System (ADS)

    Chen, Huijun; Cheng, Yuan; Liu, Dongdong; Zhang, Xiaodong; Zhang, Jue; Que, Chengli; Wang, Guangfa; Fang, Jing

    2010-03-01

    We present a structured light system to dynamically measure human chest wall motion for respiratory volume estimation. Based on a projection of an encoded color pattern and a few active markers attached to the trunk, respiratory volumes are obtained by evaluating the 3-D topographic changes of the chest wall in an anatomically consistent measuring region during respiration. Three measuring setups are established: a single-sided illuminating-recording setup for standing posture, an inclined single-sided setup for supine posture, and a double-sided setup for standing posture. Results are compared with the pneumotachography and show good agreement in volume estimations [correlation coefficient: R>0.99 (P<0.001) for all setups]. The isovolume tests present small variations of the obtained volume during the isovolume maneuver (standard deviation<0.085 L for all setups). After validation by the isovolume test, an investigation of a patient with pleural effusion using the proposed method shows pulmonary functional differences between the diseased and the contralateral sides of the thorax, and subsequent improvement of this imbalance after drainage. These results demonstrate the proposed optical method is capable of not only whole respiratory volume evaluation with high accuracy, but also regional pulmonary function assessment in different chest wall behaviors, with the advantage of whole-field measurement.

  16. Immobilization precision of a modified GTC frame

    PubMed Central

    Daartz, Juliane; Dankers, Frank; Bussière, Marc

    2012-01-01

    The purpose of this study was to evaluate and quantify the interfraction reproducibility and intrafraction immobilization precision of a modified GTC frame. The error of the patient alignment and imaging systems were measured using a cranial skull phantom, with simulated, predetermined shifts. The kV setup images were acquired with a room‐mounted set of kV sources and panels. Calculated translations and rotations provided by the computer alignment software relying upon three implanted fiducials were compared to the known shifts, and the accuracy of the imaging and positioning systems was calculated. Orthogonal kV setup images for 45 proton SRT patients and 1002 fractions (average 22.3 fractions/patient) were analyzed for interfraction and intrafraction immobilization precision using a modified GTC frame. The modified frame employs a radiotransparent carbon cup and molded pillow to allow for more treatment angles from posterior directions for cranial lesions. Patients and the phantom were aligned with three 1.5 mm stainless steel fiducials implanted into the skull. The accuracy and variance of the patient positioning and imaging systems were measured to be 0.10±0.06 mm, with the maximum uncertainty of rotation being ±0.07°.957 pairs of interfraction image sets and 974 intrafraction image sets were analyzed. 3D translations and rotations were recorded. The 3D vector interfraction setup reproducibility was 0.13 mm ±1.8 mm for translations and the largest uncertainty of ±1.07° for rotations. The intrafraction immobilization efficacy was 0.19 mm ±0.66 mm for translations and the largest uncertainty of ±0.50° for rotations. The modified GTC frame provides reproducible setup and effective intrafraction immobilization, while allowing for the complete range of entrance angles from the posterior direction. PACS number: 87.53.Ly, 87.55.Qr PMID:22584167

  17. Feasibility of an online adaptive replanning method for cranial frameless intensity-modulated radiosurgery

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Calvo, Juan Francisco, E-mail: jfcdrr@gmail.com; San José, Sol; Garrido, LLuís

    2013-10-01

    To introduce an approach for online adaptive replanning (i.e., dose-guided radiosurgery) in frameless stereotactic radiosurgery, when a 6-dimensional (6D) robotic couch is not available in the linear accelerator (linac). Cranial radiosurgical treatments are planned in our department using intensity-modulated technique. Patients are immobilized using thermoplastic mask. A cone-beam computed tomography (CBCT) scan is acquired after the initial laser-based patient setup (CBCT{sub setup}). The online adaptive replanning procedure we propose consists of a 6D registration-based mapping of the reference plan onto actual CBCT{sub setup}, followed by a reoptimization of the beam fluences (“6D plan”) to achieve similar dosage as originally wasmore » intended, while the patient is lying in the linac couch and the original beam arrangement is kept. The goodness of the online adaptive method proposed was retrospectively analyzed for 16 patients with 35 targets treated with CBCT-based frameless intensity modulated technique. Simulation of reference plan onto actual CBCT{sub setup}, according to the 4 degrees of freedom, supported by linac couch was also generated for each case (4D plan). Target coverage (D99%) and conformity index values of 6D and 4D plans were compared with the corresponding values of the reference plans. Although the 4D-based approach does not always assure the target coverage (D99% between 72% and 103%), the proposed online adaptive method gave a perfect coverage in all cases analyzed as well as a similar conformity index value as was planned. Dose-guided radiosurgery approach is effective to assure the dose coverage and conformity of an intracranial target volume, avoiding resetting the patient inside the mask in a “trial and error” way so as to remove the pitch and roll errors when a robotic table is not available.« less

  18. A system for EPID-based real-time treatment delivery verification during dynamic IMRT treatment.

    PubMed

    Fuangrod, Todsaporn; Woodruff, Henry C; van Uytven, Eric; McCurdy, Boyd M C; Kuncic, Zdenka; O'Connor, Daryl J; Greer, Peter B

    2013-09-01

    To design and develop a real-time electronic portal imaging device (EPID)-based delivery verification system for dynamic intensity modulated radiation therapy (IMRT) which enables detection of gross treatment delivery errors before delivery of substantial radiation to the patient. The system utilizes a comprehensive physics-based model to generate a series of predicted transit EPID image frames as a reference dataset and compares these to measured EPID frames acquired during treatment. The two datasets are using MLC aperture comparison and cumulative signal checking techniques. The system operation in real-time was simulated offline using previously acquired images for 19 IMRT patient deliveries with both frame-by-frame comparison and cumulative frame comparison. Simulated error case studies were used to demonstrate the system sensitivity and performance. The accuracy of the synchronization method was shown to agree within two control points which corresponds to approximately ∼1% of the total MU to be delivered for dynamic IMRT. The system achieved mean real-time gamma results for frame-by-frame analysis of 86.6% and 89.0% for 3%, 3 mm and 4%, 4 mm criteria, respectively, and 97.9% and 98.6% for cumulative gamma analysis. The system can detect a 10% MU error using 3%, 3 mm criteria within approximately 10 s. The EPID-based real-time delivery verification system successfully detected simulated gross errors introduced into patient plan deliveries in near real-time (within 0.1 s). A real-time radiation delivery verification system for dynamic IMRT has been demonstrated that is designed to prevent major mistreatments in modern radiation therapy.

  19. Fingerprint verification on medical image reporting system.

    PubMed

    Chen, Yen-Cheng; Chen, Liang-Kuang; Tsai, Ming-Dar; Chiu, Hou-Chang; Chiu, Jainn-Shiun; Chong, Chee-Fah

    2008-03-01

    The healthcare industry is recently going through extensive changes, through adoption of robust, interoperable healthcare information technology by means of electronic medical records (EMR). However, a major concern of EMR is adequate confidentiality of the individual records being managed electronically. Multiple access points over an open network like the Internet increases possible patient data interception. The obligation is on healthcare providers to procure information security solutions that do not hamper patient care while still providing the confidentiality of patient information. Medical images are also part of the EMR which need to be protected from unauthorized users. This study integrates the techniques of fingerprint verification, DICOM object, digital signature and digital envelope in order to ensure that access to the hospital Picture Archiving and Communication System (PACS) or radiology information system (RIS) is only by certified parties.

  20. Comparison of fingerprint and facial biometric verification technologies for user access and patient identification in a clinical environment

    NASA Astrophysics Data System (ADS)

    Guo, Bing; Zhang, Yu; Documet, Jorge; Liu, Brent; Lee, Jasper; Shrestha, Rasu; Wang, Kevin; Huang, H. K.

    2007-03-01

    As clinical imaging and informatics systems continue to integrate the healthcare enterprise, the need to prevent patient mis-identification and unauthorized access to clinical data becomes more apparent especially under the Health Insurance Portability and Accountability Act (HIPAA) mandate. Last year, we presented a system to track and verify patients and staff within a clinical environment. This year, we further address the biometric verification component in order to determine which Biometric system is the optimal solution for given applications in the complex clinical environment. We install two biometric identification systems including fingerprint and facial recognition systems at an outpatient imaging facility, Healthcare Consultation Center II (HCCII). We evaluated each solution and documented the advantages and pitfalls of each biometric technology in this clinical environment.

  1. Dose-response characteristics of an amorphous silicon EPID.

    PubMed

    Winkler, Peter; Hefner, Alfred; Georg, Dietmar

    2005-10-01

    Electronic portal imaging devices (EPIDs) were originally developed for the purpose of patient setup verification. Nowadays, they are increasingly used as dosimeters (e.g., for IMRT verification and linac-specific QA). A prerequisite for any clinical dosimetric application is a detailed understanding of the detector's dose-response behavior. The aim of this study is to investigate the dosimetric properties of an amorphous silicon EPID (Elekta IVIEWGT) with respect to three photon beam qualities: 6, 10, and 25 MV. The EPID showed an excellent temporal stability on short term as well as on long term scales. The stability throughout the day was strongly influenced by warming up, which took several hours and affected EPID response by 2.5%. Ghosting effects increased the sensitivity of the EPID. They became more pronounced with decreasing time intervals between two exposures as well as with increasing dose. Due to ghosting, changes in pixel sensitivity amounted up to 16% (locally) for the 25 MV photon beam. It was observed that the response characteristics of our EPID depended on dose as well as on dose rate. Doubling the dose rate increased the EPID sensitivity by 1.5%. This behavior was successfully attributed to a dose per frame effect, i.e., a nonlinear relationship between the EPID signal and the dose which was delivered to the panel between two successive readouts. The sensitivity was found to vary up to 10% in the range of 1 to 1000 monitor units. This variation was governed by two independent effects. For low doses, the EPID signal was reduced due to the linac's changing dose rate during startup. Furthermore, the detector reading was influenced by intrabeam variations of EPID sensitivity, namely, an increase of detector response during uniform exposure. For the beam qualities which were used, the response characteristics of the EPID did not depend on energy. Differences in relative dose-response curves resulted from energy dependent temporal output characteristics of the accelerator. If ghosting is prevented from affecting the results and all dose-response effects are properly corrected for, the EPID signal becomes independent of dose rate, dose, and exposure time.

  2. Pharmacist review and its impact on Singapore nursing homes

    PubMed Central

    Chia, Hui Shan; Ho, John Aik Hui; Lim, Bernadette Daolin

    2015-01-01

    INTRODUCTION There is a high prevalence of polypharmacy and inappropriate medication use in Singapore nursing homes. This study primarily explored the benefits of pharmacist reviews in local nursing homes. The secondary aims were to review the potential cost savings gained from following the pharmacists’ recommendations and to identify the possible risks associated with polypharmacy and inappropriate medication use. METHODS A retrospective period prevalence study was performed. We analysed the pharmacotherapy problems highlighted by pharmacists in three nursing homes and the rate of acceptance of pharmacists’ recommendations. Data was collected in two phases: (a) a one-month pre-setup period, during which 480 patients were reviewed (i.e. one-time review before weekly pharmacist visits); and (b) a six-month post-setup period, during which the 480 patients were reviewed again. Pharmacotherapy problems were classified according to a clinical pharmacist recommendation taxonomy and potential risks were identified. Monthly cost savings were calculated and compared with the monthly costs of pharmacist reviews. RESULTS A total of 392 pharmacotherapy problems were identified, with pharmacist recommendations noted for each problem. Among the 392 recommendations, 236 (60.2%) were accepted. The pharmacotherapy problems were analysed for potential risks, including falls (16.0%) and constipation (13.1%). The acceptance rates were higher during the post-setup period compared to the pre-setup period (p < 0.0001). Total direct acquisition cost savings during the pre- and post-setup periods were SGD 388.30 and SGD 876.69, respectively. CONCLUSION The provision of pharmaceutical care to nursing home residents resulted in improved medication safety and quality of care. PMID:26451051

  3. Set-up uncertainties: online correction with X-ray volume imaging.

    PubMed

    Kataria, Tejinder; Abhishek, Ashu; Chadha, Pranav; Nandigam, Janardhan

    2011-01-01

    To determine interfractional three-dimensional set-up errors using X-ray volumetric imaging (XVI). Between December 2007 and August 2009, 125 patients were taken up for image-guided radiotherapy using online XVI. After matching of reference and acquired volume view images, set-up errors in three translation directions were recorded and corrected online before treatment each day. Mean displacements, population systematic (Σ), and random (σ) errors were calculated and analyzed using SPSS (v16) software. Optimum clinical target volume (CTV) to planning target volume (PTV) margin was calculated using Van Herk's (2.5Σ + 0.7 σ) and Stroom's (2Σ + 0.7 σ) formula. Patients were grouped in 4 cohorts, namely brain, head and neck, thorax, and abdomen-pelvis. The mean vector displacement recorded were 0.18 cm, 0.15 cm, 0.36 cm, and 0.35 cm for brain, head and neck, thorax, and abdomen-pelvis, respectively. Analysis of individual mean set-up errors revealed good agreement with the proposed 0.3 cm isotropic margins for brain and 0.5 cm isotropic margins for head-neck. Similarly, 0.5 cm circumferential and 1 cm craniocaudal proposed margins were in agreement with thorax and abdomen-pelvic cases. The calculated mean displacements were well within CTV-PTV margin estimates of Van Herk (90% population coverage to minimum 95% prescribed dose) and Stroom (99% target volume coverage by 95% prescribed dose). Employing these individualized margins in a particular cohort ensure comparable target coverage as described in literature, which is further improved if XVI-aided set-up error detection and correction is used before treatment.

  4. SU-E-T-659: Quantitative Evaluation of Patient Setup Accuracy of Stereotactic Radiotherapy with the Frameless 6D-ExacTrac System Using Statistical Modeling

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Keeling, V; Jin, H; Hossain, S

    2015-06-15

    Purpose: To evaluate patient setup accuracy and quantify individual and cumulative positioning uncertainties associated with different hardware and software components of the stereotactic radiotherapy (SRS/SRT) with the frameless-6D-ExacTrac system. Methods: A statistical model was used to evaluate positioning uncertainties of the different components of SRS/SRT treatment with the BrainLAB 6D-ExacTrac system using the positioning shifts of 35 patients having cranial lesions (49 total lesions treated in 1, 3, 5 fractions). All these patients were immobilized with rigid head-and-neck masks, simulated with BrainLAB-localizer and planned with iPlan treatment planning system. Infrared imaging (IR) was used initially to setup patients. Then, stereoscopicmore » x-ray images (XC) were acquired and registered to corresponding digitally-reconstructed-radiographs using bony-anatomy matching to calculate 6D-translational and rotational shifts. When the shifts were within tolerance (0.7mm and 1°), treatment was initiated. Otherwise corrections were applied and additional x-rays were acquired (XV) to verify that patient position was within tolerance. Results: The uncertainties from the mask, localizer, IR-frame, x-ray imaging, MV and kV isocentricity were quantified individually. Mask uncertainty (Translational: Lateral, Longitudinal, Vertical; Rotational: Pitch, Roll, Yaw) was the largest and varied with patients in the range (−1.05−1.50mm, −5.06–3.57mm, −5.51−3.49mm; −1.40−2.40°, −1.24−1.74°, and −2.43−1.90°) obtained from mean of XC shifts for each patient. Setup uncertainty in IR positioning (0.88,2.12,1.40mm, and 0.64,0.83,0.96°) was extracted from standard-deviation of XC. Systematic uncertainties of the localizer (−0.03,−0.01,0.03mm, and −0.03,0.00,−0.01°) and frame (0.18,0.25,−1.27mm,−0.32,0.18, and 0.47°) were extracted from means of all XV setups and mean of all XC distributions, respectively. Uncertainties in isocentricity of the MV radiotherapy machine were (0.27,0.24,0.34mm) and kV-imager (0.15,−0.4,0.21mm). Conclusion: A statistical model was developed to evaluate the individual and cumulative systematic and random uncertainties induced by the different hardware and software components of the 6D-ExacTrac-system. The immobilization mask was associated with the largest positioning uncertainty.« less

  5. The Very Specific Vortex Shedding Test on VEGA Launch Vehicle

    NASA Astrophysics Data System (ADS)

    Leofanti, Jose Luis; Fotio, Domenico; Grillenbeck, Anton; Dillinger, Stephan; Scaccia, Aldo

    2012-07-01

    When tall structures are subjected to lateral wind flow, under certain conditions, vortices are shed from alternate sides of the structure inducing periodic cross wind loads on the structure. The periodic loads, in a relatively narrow and stable frequency band, can couple with the structure’s natural frequencies. To avoid this effect the VEGA Launch System (LS) comprised a decoupling device at the launch vehicle (LV) base called Anti Vortex Shedding (AVS). During the LV-Ground Segment combined test campaign in Kourou, the LV mounted on AVS was experimentally verified, including a modal characterization test, a verification under artificial operational loads and finally tested under real wind environment. The paper gives an overview on the particular aspects of test planning, the test setup preparation inside the launch pad gantry, the test performance, test results and the conclusion for the VEGA launch system’s operational readiness.

  6. VAMOS: A pathfinder for the HAWC gamma-ray observatory

    NASA Astrophysics Data System (ADS)

    Abeysekara, A. U.; Alfaro, R.; Alvarez, C.; Álvarez, J. D.; Ángeles, F.; Arceo, R.; Arteaga-Velázquez, J. C.; Avila-Aroche, A.; Ayala Solares, H. A.; Badillo, C.; Barber, A. S.; Baughman, B. M.; Bautista-Elivar, N.; Becerra Gonzalez, J.; Belmont, E.; Benítez, E.; BenZvi, S. Y.; Berley, D.; Bernal, A.; Bonilla Rosales, M.; Braun, J.; Caballero-Lopez, R. A.; Caballero-Mora, K. S.; Cabrera, I.; Carramiñana, A.; Castañeda-Martínez, L.; Castillo, M.; Cotti, U.; Cotzomi, J.; de la Fuente, E.; De León, C.; DeYoung, T.; Diaz-Azuara, A.; Diaz-Cruz, L.; Diaz Hernandez, R.; Díaz-Vélez, J. C.; Dingus, B. L.; Dultzin, D.; DuVernois, M. A.; Ellsworth, R. W.; Fernandez, A.; Fiorino, D. W.; Fraija, N.; Galindo, A.; García-Torales, G.; Garfias, F.; González, A.; González, L. X.; González, M. M.; Goodman, J. A.; Grabski, V.; Gussert, M.; Guzmán-Cerón, C.; Hampel-Arias, Z.; Harding, J. P.; Hernández-Cervantes, L.; Hui, C. M.; Hüntemeyer, P.; Imran, A.; Iriarte, A.; Karn, P.; Kieda, D.; Kunde, G. J.; Langarica, R.; Lara, A.; Lara, G.; Lauer, R. J.; Lee, W. H.; Lennarz, D.; León Vargas, H.; Linares, E. C.; Linnemann, J. T.; Longo, M.; Luna-Garcia, R.; Marinelli, A.; Martínez, L. A.; Martínez, H.; Martínez, O.; Martínez-Castro, J.; Martos, M.; Matthews, J. A. J.; McEnery, J.; Mendoza Torres, E.; Miranda-Romagnoli, P.; Moreno, E.; Mostafá, M.; Nava, J.; Nellen, L.; Newbold, M.; Noriega-Papaqui, R.; Oceguera-Becerra, T.; Page, D. P.; Patricelli, B.; Pelayo, R.; Pérez-Pérez, E. G.; Pretz, J.; Ramírez, I.; Rentería, A.; Rivière, C.; Rosa-González, D.; Ruiz-Sala, F.; Ruiz-Velasco, E. L.; Ryan, J.; Sacahui, J. R.; Salazar, H.; Salesa, F.; Sandoval, A.; Santos, E.; Schneider, M.; Silich, S.; Sinnis, G.; Smith, A. J.; Sparks Woodle, K.; Springer, R. W.; Suarez, F.; Taboada, I.; Tepe, A.; Toale, P. A.; Tollefson, K.; Torres, I.; Tinoco, S.; Ukwatta, T. N.; Valdés Galicia, J. F.; Vanegas, P.; Vázquez, A.; Villaseñor, L.; Wall, W.; Weisgarber, T.; Westerhoff, S.; Wisher, I. G.; Wood, J.; Yodh, G. B.; Younk, P. W.; Zaborov, D.; Zepeda, A.; Zhou, H.

    2015-03-01

    VAMOS was a prototype detector built in 2011 at an altitude of 4100 m a.s.l. in the state of Puebla, Mexico. The aim of VAMOS was to finalize the design, construction techniques and data acquisition system of the HAWC observatory. HAWC is an air-shower array currently under construction at the same site of VAMOS with the purpose to study the TeV sky. The VAMOS setup included six water Cherenkov detectors and two different data acquisition systems. It was in operation between October 2011 and May 2012 with an average live time of 30%. Besides the scientific verification purposes, the eight months of data were used to obtain the results presented in this paper: the detector response to the Forbush decrease of March 2012, and the analysis of possible emission, at energies above 30 GeV, for long gamma-ray bursts GRB111016B and GRB120328B.

  7. Protection of autonomous microgrids using agent-based distributed communication

    DOE PAGES

    Cintuglu, Mehmet H.; Ma, Tan; Mohammed, Osama A.

    2016-04-06

    This study presents a real-time implementation of autonomous microgrid protection using agent-based distributed communication. Protection of an autonomous microgrid requires special considerations compared to large scale distribution net-works due to the presence of power converters and relatively low inertia. In this work, we introduce a practical overcurrent and a frequency selectivity method to overcome conventional limitations. The proposed overcurrent scheme defines a selectivity mechanism considering the remedial action scheme (RAS) of the microgrid after a fault instant based on feeder characteristics and the location of the intelligent electronic devices (IEDs). A synchrophasor-based online frequency selectivity approach is proposed to avoidmore » pulse loading effects in low inertia microgrids. Experimental results are presented for verification of the pro-posed schemes using a laboratory based microgrid. The setup was composed of actual generation units and IEDs using IEC 61850 protocol. The experimental results were in excellent agreement with the proposed protection scheme.« less

  8. Automated feature detection and identification in digital point-ordered signals

    DOEpatents

    Oppenlander, Jane E.; Loomis, Kent C.; Brudnoy, David M.; Levy, Arthur J.

    1998-01-01

    A computer-based automated method to detect and identify features in digital point-ordered signals. The method is used for processing of non-destructive test signals, such as eddy current signals obtained from calibration standards. The signals are first automatically processed to remove noise and to determine a baseline. Next, features are detected in the signals using mathematical morphology filters. Finally, verification of the features is made using an expert system of pattern recognition methods and geometric criteria. The method has the advantage that standard features can be, located without prior knowledge of the number or sequence of the features. Further advantages are that standard features can be differentiated from irrelevant signal features such as noise, and detected features are automatically verified by parameters extracted from the signals. The method proceeds fully automatically without initial operator set-up and without subjective operator feature judgement.

  9. Novel Hyperspectral Sun Photometer for Satellite Remote Sensing Data Radiometric Calibration and Atmospheric Aerosol Studies

    NASA Technical Reports Server (NTRS)

    Pagnutti, Mary; Ryan, Robert E.; Holekamp, Kara; Harrington, Gary; Frisbie, Troy

    2006-01-01

    A simple and cost-effective, hyperspectral sun photometer for radiometric vicarious remote sensing system calibration, air quality monitoring, and potentially in-situ planetary climatological studies, was developed. The device was constructed solely from off the shelf components and was designed to be easily deployable for support of short-term verification and validation data collects. This sun photometer not only provides the same data products as existing multi-band sun photometers, this device requires a simpler setup, less data acquisition time and allows for a more direct calibration approach. Fielding this instrument has also enabled Stennis Space Center (SSC) Applied Sciences Directorate personnel to cross calibrate existing sun photometers. This innovative research will position SSC personnel to perform air quality assessments in support of the NASA Applied Sciences Program's National Applications program element as well as to develop techniques to evaluate aerosols in a Martian or other planetary atmosphere.

  10. Plant growth using EMCS hardware on the ISS.

    PubMed

    Iversen, Tor-Henning; Fossum, Knut R; Svare, Hakon; Johnsson, Anders; Schiller, Peter

    2002-07-01

    Under separate contracts with ESA (FUMO and ERM Study) and as a link in the development of the European Modular Cultivation System's (EMCS) functionality and biocompatibility, plant studies have been performed at The Plant Biocentre in Trondheim, Norway. The main goal was to test whether the breadboards containing the major components planned for use in the EMCS would be optimal for space experiments with plant material. The test plans and the experimental set-up for the verification of biocompatibility and biological functionality included the use of a few model plant species including cress (Lepidium sativum L.) and Arabidopsis thaliana. The plants were tested at different developmental levels of morphological and physiological complexity (illumination, life support, humidity control, water supply, observation, short- and long-term plant growth experiments and contamination prevention). Results from the tests show that the EMCS concept is useful for long duration plant growth on the ISS.

  11. In-line phase contrast micro-CT reconstruction for biomedical specimens.

    PubMed

    Fu, Jian; Tan, Renbo

    2014-01-01

    X-ray phase contrast micro computed tomography (micro-CT) can non-destructively provide the internal structure information of soft tissues and low atomic number materials. It has become an invaluable analysis tool for biomedical specimens. Here an in-line phase contrast micro-CT reconstruction technique is reported, which consists of a projection extraction method and the conventional filter back-projection (FBP) reconstruction algorithm. The projection extraction is implemented by applying the Fourier transform to the forward projections of in-line phase contrast micro-CT. This work comprises a numerical study of the method and its experimental verification using a biomedical specimen dataset measured at an X-ray tube source micro-CT setup. The numerical and experimental results demonstrate that the presented technique can improve the imaging contrast of biomedical specimens. It will be of interest for a wide range of in-line phase contrast micro-CT applications in medicine and biology.

  12. Protection of autonomous microgrids using agent-based distributed communication

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Cintuglu, Mehmet H.; Ma, Tan; Mohammed, Osama A.

    This study presents a real-time implementation of autonomous microgrid protection using agent-based distributed communication. Protection of an autonomous microgrid requires special considerations compared to large scale distribution net-works due to the presence of power converters and relatively low inertia. In this work, we introduce a practical overcurrent and a frequency selectivity method to overcome conventional limitations. The proposed overcurrent scheme defines a selectivity mechanism considering the remedial action scheme (RAS) of the microgrid after a fault instant based on feeder characteristics and the location of the intelligent electronic devices (IEDs). A synchrophasor-based online frequency selectivity approach is proposed to avoidmore » pulse loading effects in low inertia microgrids. Experimental results are presented for verification of the pro-posed schemes using a laboratory based microgrid. The setup was composed of actual generation units and IEDs using IEC 61850 protocol. The experimental results were in excellent agreement with the proposed protection scheme.« less

  13. A review of recent developments in flight test techniques at the Ames Research Center, Dryden Flight Research Facility

    NASA Technical Reports Server (NTRS)

    Layton, G. P.

    1984-01-01

    New flight test techniques in use at Ames Dryden are reviewed. The use of the pilot in combination with ground and airborne computational capabilities to maximize data return is discussed, including the remotely piloted research vehicle technique for high-risk testing, the remotely augmented vehicle technique for handling qualities research, and use of ground computed flight director information to fly unique profiles such as constant Reynolds number profiles through the transonic flight regime. Techniques used for checkout and design verification of systems-oriented aircraft are discussed, including descriptions of the various simulations, iron bird setups, and vehicle tests. Some newly developed techniques to support the aeronautical research disciplines are discussed, including a new approach to position-error determination, and the use of a large skin friction balance for the measurement of drag caused by various excrescencies.

  14. Streaming and particle motion in acoustically-actuated leaky systems

    NASA Astrophysics Data System (ADS)

    Nama, Nitesh; Barnkob, Rune; Jun Huang, Tony; Kahler, Christian; Costanzo, Francesco

    2017-11-01

    The integration of acoustics with microfluidics has shown great promise for applications within biology, chemistry, and medicine. A commonly employed system to achieve this integration consists of a fluid-filled, polymer-walled microchannel that is acoustically actuated via standing surface acoustic waves. However, despite significant experimental advancements, the precise physical understanding of such systems remains a work in progress. In this work, we investigate the nature of acoustic fields that are setup inside the microchannel as well as the fundamental driving mechanism governing the fluid and particle motion in these systems. We provide an experimental benchmark using state-of-art 3D measurements of fluid and particle motion and present a Lagrangian velocity based temporal multiscale numerical framework to explain the experimental observations. Following verification and validation, we employ our numerical model to reveal the presence of a pseudo-standing acoustic wave that drives the acoustic streaming and particle motion in these systems.

  15. Patient motion tracking in the presence of measurement errors.

    PubMed

    Haidegger, Tamás; Benyó, Zoltán; Kazanzides, Peter

    2009-01-01

    The primary aim of computer-integrated surgical systems is to provide physicians with superior surgical tools for better patient outcome. Robotic technology is capable of both minimally invasive surgery and microsurgery, offering remarkable advantages for the surgeon and the patient. Current systems allow for sub-millimeter intraoperative spatial positioning, however certain limitations still remain. Measurement noise and unintended changes in the operating room environment can result in major errors. Positioning errors are a significant danger to patients in procedures involving robots and other automated devices. We have developed a new robotic system at the Johns Hopkins University to support cranial drilling in neurosurgery procedures. The robot provides advanced visualization and safety features. The generic algorithm described in this paper allows for automated compensation of patient motion through optical tracking and Kalman filtering. When applied to the neurosurgery setup, preliminary results show that it is possible to identify patient motion within 700 ms, and apply the appropriate compensation with an average of 1.24 mm positioning error after 2 s of setup time.

  16. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Cho, M; Kim, T; Kang, S

    Purpose: The purpose of this work is to develop a new patient set-up monitoring system using force sensing resistor (FSR) sensors that can confirm pressure of contact surface and evaluate its feasibility. Methods: In this study, we focused on develop the patient set-up monitoring system to compensate for the limitation of existing optical based monitoring system, so the developed system can inform motion in the radiation therapy. The set-up monitoring system was designed consisting of sensor units (FSR sensor), signal conditioning devices (USB cable/interface electronics), a control PC, and a developed analysis software. The sensor unit was made by attachingmore » FSR sensor and dispersing pressure sponge to prevent error which is caused by concentrating specific point. Measured signal from the FSR sensor was sampled to arduino mega 2560 microcontroller, transferred to control PC by using serial communication. The measured data went through normalization process. The normalized data was displayed through the developed graphic user interface (GUI) software. The software was designed to display a single sensor unit intensity (maximum 16 sensors) and display 2D pressure distribution (using 16 sensors) according to the purpose. Results: Changes of pressure value according to motion was confirmed by the developed set-up monitoring system. Very small movement such as little physical change in appearance can be confirmed using a single unit and using 2D pressure distribution. Also, the set-up monitoring system can observe in real time. Conclusion: In this study, we developed the new set-up monitoring system using FSR sensor. Especially, we expect that the new set-up monitoring system is suitable for motion monitoring of blind area that is hard to confirm existing optical system and compensate existing optical based monitoring system. As a further study, an integrated system will be constructed through correlation of existing optical monitoring system. This work was supported by the Industrial R&D program of MOTIE/KEIT. [10048997, Development of the core technology for integrated therapy devices based on real-time MRI guided tumor tracking] and the Mid-career Researcher Program (2014R1A2A1A10050270) through the National Research Foundation of Korea funded by the Ministry of Science, ICT&Future Planning.« less

  17. Quality assurance for a six degrees-of-freedom table using a 3D printed phantom.

    PubMed

    Woods, Kyle; Ayan, Ahmet S; Woollard, Jeffrey; Gupta, Nilendu

    2018-01-01

    To establish a streamlined end-to-end test of a 6 degrees-of-freedom (6DoF) robotic table using a 3D printed phantom for periodic quality assurance. A 3D printed phantom was fabricated with translational and rotational offsets and an imbedded central ball-bearing (BB). The phantom underwent each step of the radiation therapy process: CT simulation in a straight orientation, plan generation using the treatment planning software, setup to offset marks at the linac, registration and corrected 6DoF table adjustments via hidden target test, delivery of a Winston-Lutz test to the BB, and verification of table positioning via field and laser lights. The registration values, maximum total displacement of the combined Winston-Lutz fields, and a pass or fail criterion of the laser and field lights were recorded. The quality assurance process for each of the three linacs were performed for the first 30 days. Within a 95% confidence interval, the overall uncertainty values for both translation and rotation were below 1.0 mm and 0.5° for each linac respectively. When combining the registration values and other uncertainties for all three linacs, the average deviations were within 2.0 mm and 1.0° of the designed translation and rotation offsets of the 3D print respectively. For all three linacs, the maximum total deviation for the Winston-Lutz test did not exceed 1.0 mm. Laser and light field verification was within tolerance every day for all three linacs given the latest guidance documentation for table repositioning. The 3D printer is capable of accurately fabricating a quality assurance phantom for 6DoF positioning verification. The end-to-end workflow allows for a more efficient test of the 6DoF mechanics while including other important tests needed for routine quality assurance. © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  18. A service evaluation of on-line image-guided radiotherapy to lower extremity sarcoma: Investigating the workload implications of a 3 mm action level for image assessment and correction prior to delivery.

    PubMed

    Taylor, C; Parker, J; Stratford, J; Warren, M

    2018-05-01

    Although all systematic and random positional setup errors can be corrected for in entirety during on-line image-guided radiotherapy, the use of a specified action level, below which no correction occurs, is also an option. The following service evaluation aimed to investigate the use of this 3 mm action level for on-line image assessment and correction (online, systematic set-up error and weekly evaluation) for lower extremity sarcoma, and understand the impact on imaging frequency and patient positioning error within one cancer centre. All patients were immobilised using a thermoplastic shell attached to a plastic base and an individual moulded footrest. A retrospective analysis of 30 patients was performed. Patient setup and correctional data derived from cone beam CT analysis was retrieved. The timing, frequency and magnitude of corrections were evaluated. The population systematic and random error was derived. 20% of patients had no systematic corrections over the duration of treatment, and 47% had one. The maximum number of systematic corrections per course of radiotherapy was 4, which occurred for 2 patients. 34% of episodes occurred within the first 5 fractions. All patients had at least one observed translational error during their treatment greater than 0.3 cm, and 80% of patients had at least one observed translational error during their treatment greater than 0.5 cm. The population systematic error was 0.14 cm, 0.10 cm, 0.14 cm and random error was 0.27 cm, 0.22 cm, 0.23 cm in the lateral, caudocranial and anteroposterial directions. The required Planning Target Volume margin for the study population was 0.55 cm, 0.41 cm and 0.50 cm in the lateral, caudocranial and anteroposterial directions. The 3 mm action level for image assessment and correction prior to delivery reduced the imaging burden and focussed intervention on patients that exhibited greater positional variability. This strategy could be an efficient deployment of departmental resources if full daily correction of positional setup error is not possible. Copyright © 2017. Published by Elsevier Ltd.

  19. Image-Guided Radiotherapy for Left-Sided Breast Cancer Patients: Geometrical Uncertainty of the Heart

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Topolnjak, Rajko; Borst, Gerben R.; Nijkamp, Jasper

    Purpose: To quantify the geometrical uncertainties for the heart during radiotherapy treatment of left-sided breast cancer patients and to determine and validate planning organ at risk volume (PRV) margins. Methods and Materials: Twenty-two patients treated in supine position in 28 fractions with regularly acquired cone-beam computed tomography (CBCT) scans for offline setup correction were included. Retrospectively, the CBCT scans were reconstructed into 10-phase respiration correlated four-dimensional scans. The heart was registered in each breathing phase to the planning CT scan to establish the respiratory heart motion during the CBCT scan ({sigma}{sub resp}). The average of the respiratory motion was calculatedmore » as the heart displacement error for a fraction. Subsequently, the systematic ({Sigma}), random ({sigma}), and total random ({sigma}{sub tot}={radical}({sigma}{sup 2}+{sigma}{sub resp}{sup 2})) errors of the heart position were calculated. Based on the errors a PRV margin for the heart was calculated to ensure that the maximum heart dose (D{sub max}) is not underestimated in at least 90% of the cases (M{sub heart} = 1.3{Sigma}-0.5{sigma}{sub tot}). All analysis were performed in left-right (LR), craniocaudal (CC), and anteroposterior (AP) directions with respect to both online and offline bony anatomy setup corrections. The PRV margin was validated by accumulating the dose to the heart based on the heart registrations and comparing the planned PRV D{sub max} to the accumulated heart D{sub max}. Results: For online setup correction, the cardiac geometrical uncertainties and PRV margins were N-Ary-Summation = 2.2/3.2/2.1 mm, {sigma} = 2.1/2.9/1.4 mm, and M{sub heart} = 1.6/2.3/1.3 mm for LR/CC/AP, respectively. For offline setup correction these were N-Ary-Summation = 2.4/3.7/2.2 mm, {sigma} = 2.9/4.1/2.7 mm, and M{sub heart} = 1.6/2.1/1.4 mm. Cardiac motion induced by breathing was {sigma}{sub resp} = 1.4/2.9/1.4 mm for LR/CC/AP. The PRV D{sub max} underestimated the accumulated heart D{sub max} for 9.1% patients using online and 13.6% patients using offline bony anatomy setup correction, which validated that PRV margin size was adequate. Conclusion: Considerable cardiac position variability relative to the bony anatomy was observed in breast cancer patients. A PRV margin can be used during treatment planning to take these uncertainties into account.« less

  20. SU-E-J-117: Verification Method for the Detection Accuracy of Automatic Winston Lutz Test

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Tang, A; Chan, K; Fee, F

    2014-06-01

    Purpose: Winston Lutz test (WLT) has been a standard QA procedure performed prior to SRS treatment, to verify the mechanical iso-center setup accuracy upon different Gantry/Couch movements. Several detection algorithms exist,for analyzing the ball-radiation field alignment automatically. However, the accuracy of these algorithms have not been fully addressed. Here, we reveal the possible errors arise from each step in WLT, and verify the software detection accuracy with the Rectilinear Phantom Pointer (RLPP), a tool commonly used for aligning treatment plan coordinate with mechanical iso-center. Methods: WLT was performed with the radio-opaque ball mounted on a MIS and irradiated onto EDR2more » films. The films were scanned and processed with an in-house Matlab program for automatic iso-center detection. Tests were also performed to identify the errors arise from setup, film development and scanning process. The radioopaque ball was then mounted onto the RLPP, and offset laterally and longitudinally in 7 known positions ( 0, ±0.2, ±0.5, ±0.8 mm) manually for irradiations. The gantry and couch was set to zero degree for all irradiation. The same scanned images were processed repeatedly to check the repeatability of the software. Results: Miminal discrepancies (mean=0.05mm) were detected with 2 films overlapped and irradiated but developed separately. This reveals the error arise from film processor and scanner alone. Maximum setup errors were found to be around 0.2mm, by analyzing data collected from 10 irradiations over 2 months. For the known shift introduced using the RLPP, the results agree with the manual offset, and fit linearly (R{sup 2}>0.99) when plotted relative to the first ball with zero shift. Conclusion: We systematically reveal the possible errors arise from each step in WLT, and introduce a simple method to verify the detection accuracy of our in-house software using a clinically available tool.« less

  1. SU-E-T-282: Dose Measurements with An End-To-End Audit Phantom for Stereotactic Radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jones, R; Artschan, R; Thwaites, D

    Purpose: Report on dose measurements as part of an end-to-end test for stereotactic radiotherapy, using a new audit tool, which allows audits to be performed efficiently either by an onsite team or as a postal audit. Methods: Film measurements have been performed with a new Stereotactic Cube Phantom. The phantom has been designed to perform Winston Lutz type position verification measurements and dose measurements in one setup. It comprises a plastic cube with a high density ball in its centre (used for MV imaging with film or EPID) and low density markers in the periphery (used for Cone Beam Computedmore » Tomography, CBCT imaging). It also features strategically placed gold markers near the posterior and right surfaces, which can be used to calculate phantom rotations on MV images. Slit-like openings allow insertion of film or other detectors.The phantom was scanned and small field treatment plans were created. The fields do not traverse any inhomogeneities of the phantom on their paths to the measurement location. The phantom was setup at the delivery system using CBCT imaging. The calculated treatment fields were delivered, each with a piece of radiochromic film (EBT3) placed in the anterior film holder of the phantom. MU had been selected in planning to achieve similar exposures on all films. Calibration films were exposed in solid water for dose levels around the expected doses. Films were scanned and analysed following established procedures. Results: Setup of the cube showed excellent suitability for CBCT 3D alignment. MV imaging with EPID allowed for clear identification of all markers. Film based dose measurements showed good agreement for MLC created fields down to 0.5 mm × 0.5 mm. Conclusion: An end-to-end audit phantom for stereotactic radiotherapy has been developed and tested.« less

  2. A Design of a Surgical Site Verification System.

    PubMed

    Shen, Biyu; He, Yan; Chen, Haoyang

    2017-01-01

    Patient security is a significant issue in medical research and clinical practice at present. The Surgical Verification System (Patent Number: ZL 201420079273.5) is designed to recognize and check surgical sites of patients so as to ensure operation security and decrease the risk for practitioners. Composition: (1) Operating Room Server, (2) Label Reader, (3) E-Label, (4) Surgical Site Display, (5) Ward Client, (6) Label Rader-Writer, and (7) Acousto-Optic Alarm. If the Surgical identification, the surgical site, and so on are incorrect, a flashing label control will appear when the alarm rings. You can specify a sound to play for the alarm, a picture to draw, and a message to send. It is a user-friendly system.

  3. Verification of Bioanalytical Method for Quantification of Exogenous Insulin (Insulin Aspart) by the Analyser Advia Centaur® XP.

    PubMed

    Mihailov, Rossen; Stoeva, Dilyana; Pencheva, Blagovesta; Pentchev, Eugeni

    2018-03-01

    In a number of cases the monitoring of patients with type I diabetes mellitus requires measurement of the exogenous insulin levels. For the purpose of a clinical investigation of the efficacy of a medical device for application of exogenous insulin aspart, a verification of the method for measurement of this synthetic analogue of the hormone was needed. The information in the available medical literature for the measurement of the different exogenous insulin analogs is insufficient. Thus, verification was required to be in compliance with the active standards in Republic of Bulgaria. A manufactured method developed for ADVIA Centaur XP Immunoassay, Siemens Healthcare, was used which we verified using standard solutions and a patient serum pool by adding the appropriate quantity exogenous insulin aspart. The method was verified in accordance with the bioanalytical method verification criteria and regulatory requirements for using a standard method: CLIA chemiluminescence immunoassay ADVIA Centaur® XP. The following parameters are determined and monitored: intra-day precision and accuracy, inter-day precision and accuracy, limit of detection and lower limit of quantification, linearity, analytical recovery. The routine application of the method for measurement of immunoreactive insulin using the analyzer ADVIA Centaur® XP is directed to the measurement of endogenous insulin. The method is applicable for measuring different types of exogenous insulin, including insulin aspart.

  4. Calibrating page sized Gafchromic EBT3 films

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Crijns, W.; Maes, F.; Heide, U. A. van der

    2013-01-15

    Purpose: The purpose is the development of a novel calibration method for dosimetry with Gafchromic EBT3 films. The method should be applicable for pretreatment verification of volumetric modulated arc, and intensity modulated radiotherapy. Because the exposed area on film can be large for such treatments, lateral scan errors must be taken into account. The correction for the lateral scan effect is obtained from the calibration data itself. Methods: In this work, the film measurements were modeled using their relative scan values (Transmittance, T). Inside the transmittance domain a linear combination and a parabolic lateral scan correction described the observed transmittancemore » values. The linear combination model, combined a monomer transmittance state (T{sub 0}) and a polymer transmittance state (T{sub {infinity}}) of the film. The dose domain was associated with the observed effects in the transmittance domain through a rational calibration function. On the calibration film only simple static fields were applied and page sized films were used for calibration and measurements (treatment verification). Four different calibration setups were considered and compared with respect to dose estimation accuracy. The first (I) used a calibration table from 32 regions of interest (ROIs) spread on 4 calibration films, the second (II) used 16 ROIs spread on 2 calibration films, the third (III), and fourth (IV) used 8 ROIs spread on a single calibration film. The calibration tables of the setups I, II, and IV contained eight dose levels delivered to different positions on the films, while for setup III only four dose levels were applied. Validation was performed by irradiating film strips with known doses at two different time points over the course of a week. Accuracy of the dose response and the lateral effect correction was estimated using the dose difference and the root mean squared error (RMSE), respectively. Results: A calibration based on two films was the optimal balance between cost effectiveness and dosimetric accuracy. The validation resulted in dose errors of 1%-2% for the two different time points, with a maximal absolute dose error around 0.05 Gy. The lateral correction reduced the RMSE values on the sides of the film to the RMSE values at the center of the film. Conclusions: EBT3 Gafchromic films were calibrated for large field dosimetry with a limited number of page sized films and simple static calibration fields. The transmittance was modeled as a linear combination of two transmittance states, and associated with dose using a rational calibration function. Additionally, the lateral scan effect was resolved in the calibration function itself. This allows the use of page sized films. Only two calibration films were required to estimate both the dose and the lateral response. The calibration films were used over the course of a week, with residual dose errors Less-Than-Or-Slanted-Equal-To 2% or Less-Than-Or-Slanted-Equal-To 0.05 Gy.« less

  5. Development of a Targeted Smoking Relapse-Prevention Intervention for Cancer Patients.

    PubMed

    Meltzer, Lauren R; Meade, Cathy D; Diaz, Diana B; Carrington, Monica S; Brandon, Thomas H; Jacobsen, Paul B; McCaffrey, Judith C; Haura, Eric B; Simmons, Vani N

    2018-04-01

    We describe the series of iterative steps used to develop a smoking relapse-prevention intervention customized to the needs of cancer patients. Informed by relevant literature and a series of preliminary studies, an educational tool (DVD) was developed to target the unique smoking relapse risk factors among cancer patients. Learner verification interviews were conducted with 10 cancer patients who recently quit smoking to elicit feedback and inform the development of the DVD. The DVD was then refined using iterative processes and feedback from the learner verification interviews. Major changes focused on visual appeal, and the inclusion of additional testimonials and graphics to increase comprehension of key points and further emphasize the message that the patient is in control of their ability to maintain their smoking abstinence. Together, these steps resulted in the creation of a DVD titled Surviving Smokefree®, which represents the first smoking relapse-prevention intervention for cancer patients. If found effective, the Surviving Smokefree® DVD is an easily disseminable and low-cost portable intervention which can assist cancer patients in maintaining smoking abstinence.

  6. Image guidance during head-and-neck cancer radiation therapy: analysis of alignment trends with in-room cone-beam computed tomography scans.

    PubMed

    Zumsteg, Zachary; DeMarco, John; Lee, Steve P; Steinberg, Michael L; Lin, Chun Shu; McBride, William; Lin, Kevin; Wang, Pin-Chieh; Kupelian, Patrick; Lee, Percy

    2012-06-01

    On-board cone-beam computed tomography (CBCT) is currently available for alignment of patients with head-and-neck cancer before radiotherapy. However, daily CBCT is time intensive and increases the overall radiation dose. We assessed the feasibility of using the average couch shifts from the first several CBCTs to estimate and correct for the presumed systematic setup error. 56 patients with head-and-neck cancer who received daily CBCT before intensity-modulated radiation therapy had recorded shift values in the medial-lateral, superior-inferior, and anterior-posterior dimensions. The average displacements in each direction were calculated for each patient based on the first five or 10 CBCT shifts and were presumed to represent the systematic setup error. The residual error after this correction was determined by subtracting the calculated shifts from the shifts obtained using daily CBCT. The magnitude of the average daily residual three-dimensional (3D) error was 4.8 ± 1.4 mm, 3.9 ± 1.3 mm, and 3.7 ± 1.1 mm for uncorrected, five CBCT corrected, and 10 CBCT corrected protocols, respectively. With no image guidance, 40.8% of fractions would have been >5 mm off target. Using the first five CBCT shifts to correct subsequent fractions, this percentage decreased to 19.0% of all fractions delivered and decreased the percentage of patients with average daily 3D errors >5 mm from 35.7% to 14.3% vs. no image guidance. Using an average of the first 10 CBCT shifts did not significantly improve this outcome. Using the first five CBCT shift measurements as an estimation of the systematic setup error improves daily setup accuracy for a subset of patients with head-and-neck cancer receiving intensity-modulated radiation therapy and primarily benefited those with large 3D correction vectors (>5 mm). Daily CBCT is still necessary until methods are developed that more accurately determine which patients may benefit from alternative imaging strategies. Copyright © 2012 Elsevier Inc. All rights reserved.

  7. Characterization and clinical evaluation of a novel 2D detector array for conventional and flattening filter free (FFF) IMRT pre-treatment verification.

    PubMed

    Sekar, Yuvaraj; Thoelking, Johannes; Eckl, Miriam; Kalichava, Irakli; Sihono, Dwi Seno Kuncoro; Lohr, Frank; Wenz, Frederik; Wertz, Hansjoerg

    2018-04-01

    The novel MatriXX FFF (IBA Dosimetry, Germany) detector is a new 2D ionization chamber detector array designed for patient specific IMRT-plan verification including flattening-filter-free (FFF) beams. This study provides a detailed analysis of the characterization and clinical evaluation of the new detector array. The verification of the MatriXX FFF was subdivided into (i) physical dosimetric tests including dose linearity, dose rate dependency and output factor measurements and (ii) patient specific IMRT pre-treatment plan verifications. The MatriXX FFF measurements were compared to the calculated dose distribution of a commissioned treatment planning system by gamma index and dose difference evaluations for 18 IMRT-sequences. All IMRT-sequences were measured with original gantry angles and with collapsing all beams to 0° gantry angle to exclude the influence of the detector's angle dependency. The MatriXX FFF was found to be linear and dose rate independent for all investigated modalities (deviations ≤0.6%). Furthermore, the output measurements of the MatriXX FFF were in very good agreement to reference measurements (deviations ≤1.8%). For the clinical evaluation an average pixel passing rate for γ (3%,3mm) of (98.5±1.5)% was achieved when applying a gantry angle correction. Also, with collapsing all beams to 0° gantry angle an excellent agreement to the calculated dose distribution was observed (γ (3%,3mm) =(99.1±1.1)%). The MatriXX FFF fulfills all physical requirements in terms of dosimetric accuracy. Furthermore, the evaluation of the IMRT-plan measurements showed that the detector particularly together with the gantry angle correction is a reliable device for IMRT-plan verification including FFF. Copyright © 2017. Published by Elsevier GmbH.

  8. Limited Impact of Setup and Range Uncertainties, Breathing Motion, and Interplay Effects in Robustly Optimized Intensity Modulated Proton Therapy for Stage III Non-small Cell Lung Cancer

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Inoue, Tatsuya; Widder, Joachim; Dijk, Lisanne V. van

    2016-11-01

    Purpose: To investigate the impact of setup and range uncertainties, breathing motion, and interplay effects using scanning pencil beams in robustly optimized intensity modulated proton therapy (IMPT) for stage III non-small cell lung cancer (NSCLC). Methods and Materials: Three-field IMPT plans were created using a minimax robust optimization technique for 10 NSCLC patients. The plans accounted for 5- or 7-mm setup errors with ±3% range uncertainties. The robustness of the IMPT nominal plans was evaluated considering (1) isotropic 5-mm setup errors with ±3% range uncertainties; (2) breathing motion; (3) interplay effects; and (4) a combination of items 1 and 2.more » The plans were calculated using 4-dimensional and average intensity projection computed tomography images. The target coverage (TC, volume receiving 95% of prescribed dose) and homogeneity index (D{sub 2} − D{sub 98}, where D{sub 2} and D{sub 98} are the least doses received by 2% and 98% of the volume) for the internal clinical target volume, and dose indexes for lung, esophagus, heart and spinal cord were compared with that of clinical volumetric modulated arc therapy plans. Results: The TC and homogeneity index for all plans were within clinical limits when considering the breathing motion and interplay effects independently. The setup and range uncertainties had a larger effect when considering their combined effect. The TC decreased to <98% (clinical threshold) in 3 of 10 patients for robust 5-mm evaluations. However, the TC remained >98% for robust 7-mm evaluations for all patients. The organ at risk dose parameters did not significantly vary between the respective robust 5-mm and robust 7-mm evaluations for the 4 error types. Compared with the volumetric modulated arc therapy plans, the IMPT plans showed better target homogeneity and mean lung and heart dose parameters reduced by about 40% and 60%, respectively. Conclusions: In robustly optimized IMPT for stage III NSCLC, the setup and range uncertainties, breathing motion, and interplay effects have limited impact on target coverage, dose homogeneity, and organ-at-risk dose parameters.« less

  9. Independent dose verification system with Monte Carlo simulations using TOPAS for passive scattering proton therapy at the National Cancer Center in Korea

    NASA Astrophysics Data System (ADS)

    Shin, Wook-Geun; Testa, Mauro; Kim, Hak Soo; Jeong, Jong Hwi; Byeong Lee, Se; Kim, Yeon-Joo; Min, Chul Hee

    2017-10-01

    For the independent validation of treatment plans, we developed a fully automated Monte Carlo (MC)-based patient dose calculation system with the tool for particle simulation (TOPAS) and proton therapy machine installed at the National Cancer Center in Korea to enable routine and automatic dose recalculation for each patient. The proton beam nozzle was modeled with TOPAS to simulate the therapeutic beam, and MC commissioning was performed by comparing percent depth dose with the measurement. The beam set-up based on the prescribed beam range and modulation width was automated by modifying the vendor-specific method. The CT phantom was modeled based on the DICOM CT files with TOPAS-built-in function, and an in-house-developed C++ code directly imports the CT files for positioning the CT phantom, RT-plan file for simulating the treatment plan, and RT-structure file for applying the Hounsfield unit (HU) assignment, respectively. The developed system was validated by comparing the dose distributions with those calculated by the treatment planning system (TPS) for a lung phantom and two patient cases of abdomen and internal mammary node. The results of the beam commissioning were in good agreement of up to 0.8 mm2 g-1 for B8 option in both of the beam range and the modulation width of the spread-out Bragg peaks. The beam set-up technique can predict the range and modulation width with an accuracy of 0.06% and 0.51%, respectively, with respect to the prescribed range and modulation in arbitrary points of B5 option (128.3, 132.0, and 141.2 mm2 g-1 of range). The dose distributions showed higher than 99% passing rate for the 3D gamma index (3 mm distance to agreement and 3% dose difference) between the MC simulations and the clinical TPS in the target volume. However, in the normal tissues, less favorable agreements were obtained for the radiation treatment planning with the lung phantom and internal mammary node cases. The discrepancies might come from the limitations of the clinical TPS, which is the inaccurate dose calculation algorithm for the scattering effect, in the range compensator and inhomogeneous material. Moreover, the steep slope of the compensator, conversion of the HU values to the human phantom, and the dose calculation algorithm for the HU assignment also could be reasons of the discrepancies. The current study could be used for the independent dose validation of treatment plans including high inhomogeneities, the steep compensator, and riskiness such as lung, head & neck cases. According to the treatment policy, the dose discrepancies predicted with MC could be used for the acceptance decision of the original treatment plan.

  10. SU-F-T-32: Evaluation of the Performance of a Multiple-Array-Diode Detector for Quality Assurance Tests in High-Dose-Rate Brachytherapy with Ir-192 Source

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Harpool, K; De La Fuente Herman, T; Ahmad, S

    Purpose: To evaluate the performance of a two-dimensional (2D) array-diode- detector for geometric and dosimetric quality assurance (QA) tests of high-dose-rate (HDR) brachytherapy with an Ir-192-source. Methods: A phantom setup was designed that encapsulated a two-dimensional (2D) array-diode-detector (MapCheck2) and a catheter for the HDR brachytherapy Ir-192 source. This setup was used to perform both geometric and dosimetric quality assurance for the HDR-Ir192 source. The geometric tests included: (a) measurement of the position of the source and (b) spacing between different dwell positions. The dosimteric tests include: (a) linearity of output with time, (b) end effect and (c) relative dosemore » verification. The 2D-dose distribution measured with MapCheck2 was used to perform the previous tests. The results of MapCheck2 were compared with the corresponding quality assurance testes performed with Gafchromic-film and well-ionization-chamber. Results: The position of the source and the spacing between different dwell-positions were reproducible within 1 mm accuracy by measuring the position of maximal dose using MapCheck2 in contrast to the film which showed a blurred image of the dwell positions due to limited film sensitivity to irradiation. The linearity of the dose with dwell times measured from MapCheck2 was superior to the linearity measured with ionization chamber due to higher signal-to-noise ratio of the diode readings. MapCheck2 provided more accurate measurement of the end effect with uncertainty < 1.5% in comparison with the ionization chamber uncertainty of 3%. Although MapCheck2 did not provide absolute calibration dosimeter for the activity of the source, it provided accurate tool for relative dose verification in HDR-brachytherapy. Conclusion: The 2D-array-diode-detector provides a practical, compact and accurate tool to perform quality assurance for HDR-brachytherapy with an Ir-192 source. The diodes in MapCheck2 have high radiation sensitivity and linearity that is superior to Gafchromic-films and ionization chamber used for geometric and dosimetric QA in HDR-brachytherapy, respectively.« less

  11. HVI-Test Setup for Debris Detector Verification

    NASA Astrophysics Data System (ADS)

    Bauer, Waldemar; Romberg, Oliver; Wiedemann, Carsten; Putzar, Robin; Drolshagen, Gerhard; Vorsmann, Peter

    2013-08-01

    Risk assessment concerning impacting space debris or micrometeoroids with spacecraft or payloads can be performed by using environmental models such as MASTER (ESA) or ORDEM (NASA). The validation of such models is performed by comparison of simulated results with measured data. Such data can be obtained from ground-based or space-based radars or telescopes, or by analysis of space hardware (e.g. Hubble Space Telescope, Space Shuttle Windows), which are retrieved from orbit. An additional data source is in-situ impact detectors, which are purposed for the collection of space debris and micrometeoroids impact data. In comparison to the impact data gained by analysis of the retrieved surfaces, the detected data contains additional information regarding impact time and orbit. In the past, many such in-situ detectors have been developed, with different measurement methods for the identification and classification of impacting objects. However, existing detectors have a drawback in terms of data acquisition. Generally the detection area is small, limiting the collected data as the number of recorded impacts has a linear dependence to the exposed area. An innovative impact detector concept is currently under development at the German Aerospace Centre (DLR) in Bremen, in order to increase the surface area while preserving the advantages offered by dedicated in-situ impact detectors. The Solar Generator based Impact Detector (SOLID) is not an add-on component on the spacecraft, making it different to all previous impact detectors. SOLID utilises existing subsystems of the spacecraft and adapts them for impact detection purposes. Solar generators require large panel surfaces in order to provide the spacecraft with sufficient energy. Therefore, the spacecraft solar panels provide a perfect opportunity for application as impact detectors. Employment of the SOLID method in several spacecraft in various orbits would serve to significantly increase the spatial coverage concerning space debris and micrometeoroids. In this way, the SOLID method will allow the generation of a large amount of impact data for environmental model validation. The ground verification of the SOLID method was performed at Fraunhofer EMI. For this purpose, a test model was developed. This paper focuses on the test methodology and development of the Hypervelocity Impact (HVI) test setup, including pretesting at the German Aerospace Centre (DLR), Bremen. Foreseen hardware and software for the automatic damage assessment of the detector after the impact are also presented.

  12. [Accelerated partial breast irradiation with image-guided intensity-modulated radiotherapy following breast-conserving surgery - preliminary results of a phase II clinical study].

    PubMed

    Mészáros, Norbert; Major, Tibor; Stelczer, Gábor; Zaka, Zoltán; Mózsa, Emõke; Fodor, János; Polgár, Csaba

    2015-06-01

    The purpose of the study was to implement accelerated partial breast irradiation (APBI) by means of image-guided intensity-modulated radiotherapy (IG-IMRT) following breast-conserving surgery (BCS) for low-risk early invasive breast cancer. Between July 2011 and March 2014, 60 patients with low-risk early invasive (St I-II) breast cancer who underwent BCS were enrolled in our phase II prospective study. Postoperative APBI was given by means of step and shoot IG-IMRT using 4 to 5 fields to a total dose of 36.9 Gy (9×4.1 Gy) using a twice-a-day fractionation. Before each fraction, series of CT images were taken from the region of the target volume using a kV CT on-rail mounted in the treatment room. An image fusion software was used for automatic image registration of the planning and verification CT images. Patient set-up errors were detected in three directions (LAT, LONG, VERT), and inaccuracies were adjusted by automatic movements of the treatment table. Breast cancer related events, acute and late toxicities, and cosmetic results were registered and analysed. At a median follow-up of 24 months (range 12-44) neither locoregional nor distant failure was observed. Grade 1 (G1), G2 erythema, G1 oedema, and G1 and G2 pain occurred in 21 (35%), 2 (3.3%), 23 (38.3%), 6 (10%) and 2 (3.3%) patients, respectively. No G3-4 acute side effects were detected. Among late radiation side effects G1 pigmentation, G1 fibrosis, and G1 fat necrosis occurred in 5 (8.3%), 7 (11.7%), and 2 (3.3%) patients, respectively. No ≥G2 late toxicity was detected. Excellent and good cosmetic outcome was detected in 45 (75%) and 15 (25%) patients. IG-IMRT is a reproducible and feasible technique for the delivery of APBI following conservative surgery for the treatment of low-risk, early-stage invasive breast carcinoma. Preliminary results are promising, early radiation side effects are minimal, and cosmetic results are excellent.

  13. Evaluation of Robustness to Setup and Range Uncertainties for Head and Neck Patients Treated With Pencil Beam Scanning Proton Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Malyapa, Robert; Lowe, Matthew; Christie Medical Physics and Engineering, The Christie NHS Foundation Trust, Manchester

    Purpose: To evaluate the robustness of head and neck plans for treatment with intensity modulated proton therapy to range and setup errors, and to establish robustness parameters for the planning of future head and neck treatments. Methods and Materials: Ten patients previously treated were evaluated in terms of robustness to range and setup errors. Error bar dose distributions were generated for each plan, from which several metrics were extracted and used to define a robustness database of acceptable parameters over all analyzed plans. The patients were treated in sequentially delivered series, and plans were evaluated for both the first seriesmore » and for the combined error over the whole treatment. To demonstrate the application of such a database in the head and neck, for 1 patient, an alternative treatment plan was generated using a simultaneous integrated boost (SIB) approach and plans of differing numbers of fields. Results: The robustness database for the treatment of head and neck patients is presented. In an example case, comparison of single and multiple field plans against the database show clear improvements in robustness by using multiple fields. A comparison of sequentially delivered series and an SIB approach for this patient show both to be of comparable robustness, although the SIB approach shows a slightly greater sensitivity to uncertainties. Conclusions: A robustness database was created for the treatment of head and neck patients with intensity modulated proton therapy based on previous clinical experience. This will allow the identification of future plans that may benefit from alternative planning approaches to improve robustness.« less

  14. SU-F-P-03: Management of Time to Treatment Inititation: Case for An Electronic Whiteboard

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Adnani, N

    2016-06-15

    Purpose: To determine if data mining of an electronic whiteboard improves the management of the Time to Treatment Initiation (TTI) in radiation oncology. Methods: An electronic whiteboard designed to help in managing the planning workflow and improves communication regarding patient planning progress was used to record the dates at which each phase of the planning process began or completed. These are CT Sim date, Plan Start, Physician Review, Physicist Review, Approval for Treatment Delivery, Setup or Verification of Simulation. Results: During clinical implementation, the electronic whiteboard was able to fulfill its primary objective of providing a transparent account of themore » planning progress of each patient. Peer pressure also meant that individual tasks, such as contouring, were easily brought to the attention of the responsible party and prioritized accordingly. Data mining to analyze the electronic whiteboard per patient (figure 1), per diagnosis (figure 2), per treatment modality (figure 3), per physician (figure 4), per planner (figure 5), etc., added another sophisticated tool in the management of Time to Treatment Initiation without compromising quality of the plans being generated. A longer than necessary time between CT Sim and Plan Start can be discussed among the members of the treatment team as an indication of inadequate/outdated CT Simulator, Contouring Tools, Image Fusion Tools, Other Imaging Studies (MRI, PET/CT) performed, etc. The same for the Plan Start to Physician Review where an extended time than expected may be due unrealistic planning goals, limited planning system features, etc. Conclusion: An Electronic Whiteboard in radiation oncology is not only helping with organizing planning workflow, it is also a potent tool that can be used to reduce the Time to Treatment Initiation by providing the clinic with hard data about the duration of each phase treatment planning as a function of different variable affecting the planning process. The work is supported by the Global Medical Physics Institute.« less

  15. SU-F-T-229: A Novel Method for EPID-Based In-Vivo Exit Dose Verification for Intensity Modulated Radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wu, Z; Wang, J; Peng, J

    Purpose: Electronic portal imaging device (EPID) can be used to acquire a two-dimensional exit dose distribution during treatment delivery, thus allowing the in-vivo verification of the dose delivery through a comparison of measured portal images to predicted portal dose images (PDI). The aim of this study was to present a novel method to easily and accurately predict PDI, and to establish an EPID-based in-vivo dose verification method during IMRT treatments. Methods: We developed a model to determine the predicted portal dose at the same plane of the EPID detector location. The Varian EPID (aS1000) positions at 150cm source-to-detector-distance (SDD), andmore » can be used to acquire in-vivo exit dose using Portal Dosimetry (PD) function. Our model was generated to make an equivalent water thickness represent the buildup plate of EPID. The exit dose at extend SDD plane with patient CT data in the beam can be calculated as the predicted PDI in the treatment planning system (TPS). After that, the PDI was converted to the fluence at SDD of 150cm using the inverse square law coded in MATLAB. Five head-and-neck and prostate IMRT patient plans contain 32 fields were investigated to evaluate the feasibility of this new method. The measured EPID image was compared with PDI using the gamma analysis. Results: The average results for cumulative dose comparison were 81.9% and 91.6% for 3%, 3mm and 4%, 4mm gamma criteria, respectively. Results indicate that the patient transit dosimetry predicted algorithm compares well with EPID measured PD doses for test situations. Conclusion: Our new method can be used as an easy and feasible tool for online EPID-based in-vivo dose delivery verification for IMRT treatments. It can be implemented for fast detecting those obvious treatment delivery errors for individual field and patient quality assurance.« less

  16. SU-E-T-586: Optimal Determination of Tolerance Level for Radiation Dose Delivery Verification in An in Vivo Dosimetry System

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chen, Y; Souri, S; Gill, G

    Purpose: To statistically determine the optimal tolerance level in the verification of delivery dose compared to the planned dose in an in vivo dosimetry system in radiotherapy. Methods: The LANDAUER MicroSTARii dosimetry system with screened nanoDots (optically stimulated luminescence dosimeters) was used for in vivo dose measurements. Ideally, the measured dose should match with the planned dose and falls within a normal distribution. Any deviation from the normal distribution may be redeemed as a mismatch, therefore a potential sign of the dose misadministration. Randomly mis-positioned nanoDots can yield a continuum background distribution. A percentage difference of the measured dose tomore » its corresponding planned dose (ΔD) can be used to analyze combined data sets for different patients. A model of a Gaussian plus a flat function was used to fit the ΔD distribution. Results: Total 434 nanoDot measurements for breast cancer patients were collected across a period of three months. The fit yields a Gaussian mean of 2.9% and a standard deviation (SD) of 5.3%. The observed shift of the mean from zero is attributed to the machine output bias and calibration of the dosimetry system. A pass interval of −2SD to +2SD was applied and a mismatch background was estimated to be 4.8%. With such a tolerance level, one can expect that 99.99% of patients should pass the verification and at most 0.011% might have a potential dose misadministration that may not be detected after 3 times of repeated measurements. After implementation, a number of new start breast cancer patients were monitored and the measured pass rate is consistent with the model prediction. Conclusion: It is feasible to implement an optimal tolerance level in order to maintain a low limit of potential dose misadministration while still to keep a relatively high pass rate in radiotherapy delivery verification.« less

  17. Phenotype Instance Verification and Evaluation Tool (PIVET): A Scaled Phenotype Evidence Generation Framework Using Web-Based Medical Literature.

    PubMed

    Henderson, Jette; Ke, Junyuan; Ho, Joyce C; Ghosh, Joydeep; Wallace, Byron C

    2018-05-04

    Researchers are developing methods to automatically extract clinically relevant and useful patient characteristics from raw healthcare datasets. These characteristics, often capturing essential properties of patients with common medical conditions, are called computational phenotypes. Being generated by automated or semiautomated, data-driven methods, such potential phenotypes need to be validated as clinically meaningful (or not) before they are acceptable for use in decision making. The objective of this study was to present Phenotype Instance Verification and Evaluation Tool (PIVET), a framework that uses co-occurrence analysis on an online corpus of publically available medical journal articles to build clinical relevance evidence sets for user-supplied phenotypes. PIVET adopts a conceptual framework similar to the pioneering prototype tool PheKnow-Cloud that was developed for the phenotype validation task. PIVET completely refactors each part of the PheKnow-Cloud pipeline to deliver vast improvements in speed without sacrificing the quality of the insights PheKnow-Cloud achieved. PIVET leverages indexing in NoSQL databases to efficiently generate evidence sets. Specifically, PIVET uses a succinct representation of the phenotypes that corresponds to the index on the corpus database and an optimized co-occurrence algorithm inspired by the Aho-Corasick algorithm. We compare PIVET's phenotype representation with PheKnow-Cloud's by using PheKnow-Cloud's experimental setup. In PIVET's framework, we also introduce a statistical model trained on domain expert-verified phenotypes to automatically classify phenotypes as clinically relevant or not. Additionally, we show how the classification model can be used to examine user-supplied phenotypes in an online, rather than batch, manner. PIVET maintains the discriminative power of PheKnow-Cloud in terms of identifying clinically relevant phenotypes for the same corpus with which PheKnow-Cloud was originally developed, but PIVET's analysis is an order of magnitude faster than that of PheKnow-Cloud. Not only is PIVET much faster, it can be scaled to a larger corpus and still retain speed. We evaluated multiple classification models on top of the PIVET framework and found ridge regression to perform best, realizing an average F1 score of 0.91 when predicting clinically relevant phenotypes. Our study shows that PIVET improves on the most notable existing computational tool for phenotype validation in terms of speed and automation and is comparable in terms of accuracy. ©Jette Henderson, Junyuan Ke, Joyce C Ho, Joydeep Ghosh, Byron C Wallace. Originally published in the Journal of Medical Internet Research (http://www.jmir.org), 04.05.2018.

  18. Phenotype Instance Verification and Evaluation Tool (PIVET): A Scaled Phenotype Evidence Generation Framework Using Web-Based Medical Literature

    PubMed Central

    Ke, Junyuan; Ho, Joyce C; Ghosh, Joydeep; Wallace, Byron C

    2018-01-01

    Background Researchers are developing methods to automatically extract clinically relevant and useful patient characteristics from raw healthcare datasets. These characteristics, often capturing essential properties of patients with common medical conditions, are called computational phenotypes. Being generated by automated or semiautomated, data-driven methods, such potential phenotypes need to be validated as clinically meaningful (or not) before they are acceptable for use in decision making. Objective The objective of this study was to present Phenotype Instance Verification and Evaluation Tool (PIVET), a framework that uses co-occurrence analysis on an online corpus of publically available medical journal articles to build clinical relevance evidence sets for user-supplied phenotypes. PIVET adopts a conceptual framework similar to the pioneering prototype tool PheKnow-Cloud that was developed for the phenotype validation task. PIVET completely refactors each part of the PheKnow-Cloud pipeline to deliver vast improvements in speed without sacrificing the quality of the insights PheKnow-Cloud achieved. Methods PIVET leverages indexing in NoSQL databases to efficiently generate evidence sets. Specifically, PIVET uses a succinct representation of the phenotypes that corresponds to the index on the corpus database and an optimized co-occurrence algorithm inspired by the Aho-Corasick algorithm. We compare PIVET’s phenotype representation with PheKnow-Cloud’s by using PheKnow-Cloud’s experimental setup. In PIVET’s framework, we also introduce a statistical model trained on domain expert–verified phenotypes to automatically classify phenotypes as clinically relevant or not. Additionally, we show how the classification model can be used to examine user-supplied phenotypes in an online, rather than batch, manner. Results PIVET maintains the discriminative power of PheKnow-Cloud in terms of identifying clinically relevant phenotypes for the same corpus with which PheKnow-Cloud was originally developed, but PIVET’s analysis is an order of magnitude faster than that of PheKnow-Cloud. Not only is PIVET much faster, it can be scaled to a larger corpus and still retain speed. We evaluated multiple classification models on top of the PIVET framework and found ridge regression to perform best, realizing an average F1 score of 0.91 when predicting clinically relevant phenotypes. Conclusions Our study shows that PIVET improves on the most notable existing computational tool for phenotype validation in terms of speed and automation and is comparable in terms of accuracy. PMID:29728351

  19. Impact of prospective verification of intravenous antibiotics in an ED.

    PubMed

    Hunt, Allyson; Nakajima, Steven; Hall Zimmerman, Lisa; Patel, Manav

    2016-12-01

    Delay in appropriate antibiotic therapy is associated with an increase in mortality and prolonged length of stay. Automatic dispensing machines decrease the delivery time of intravenous (IV) antibiotics to patients in the emergency department (ED). However, when IV antibiotics are not reviewed by pharmacists before being administered, patients are at risk for receiving inappropriate antibiotic therapy. The objective of this study was to determine if a difference exists in the time to administration of appropriate antibiotic therapy before and after implementation of prospective verification of antibiotics in the ED. This retrospective, institutional review board-approved preimplementation vs postimplementation study evaluated patients 18years or older who were started on IV antibiotics in the ED. Patients were excluded if pregnant, if the patient is a prisoner, if no cultures were drawn, or if the patient was transferred from an outside facility. Appropriate antibiotic therapy was based on empiric source-specific evidence-based guidelines, appropriate pharmacokinetic and pharmacodynamic properties, and microbiologic data. The primary end point was the time from ED arrival to administration of appropriate antibiotic therapy. Of the 1628 evaluated, 128 patients met the inclusion criteria (64 pre vs 64 post). Patients were aged 65.2±17.0years, with most of infections being pneumonia (44%) and urinary tract infections (18%) and most patients being noncritically ill. Time to appropriate antibiotic therapy was reduced in the postgroup vs pregroup (8.1±8.6 vs 15.2±22.8hours, respectively, P=.03). In addition, appropriate empiric antibiotics were initiated more frequently after the implementation (92% post vs 66% pre; P=.0001). There was no difference in mortality or length of stay between the 2 groups. Prompt administration of the appropriate antibiotics is imperative in patients with infections presenting to the ED. The impact of prospective verification of antibiotics by pharmacists led to significant improvement on both empiric selection of and time to appropriate antibiotic therapy. Copyright © 2016 Elsevier Inc. All rights reserved.

  20. Dosimetric accuracy of Kodak EDR2 film for IMRT verifications.

    PubMed

    Childress, Nathan L; Salehpour, Mohammad; Dong, Lei; Bloch, Charles; White, R Allen; Rosen, Isaac I

    2005-02-01

    Patient-specific intensity-modulated radiotherapy (IMRT) verifications require an accurate two-dimensional dosimeter that is not labor-intensive. We assessed the precision and reproducibility of film calibrations over time, measured the elemental composition of the film, measured the intermittency effect, and measured the dosimetric accuracy and reproducibility of calibrated Kodak EDR2 film for single-beam verifications in a solid water phantom and for full-plan verifications in a Rexolite phantom. Repeated measurements of the film sensitometric curve in a single experiment yielded overall uncertainties in dose of 2.1% local and 0.8% relative to 300 cGy. 547 film calibrations over an 18-month period, exposed to a range of doses from 0 to a maximum of 240 MU or 360 MU and using 6 MV or 18 MV energies, had optical density (OD) standard deviations that were 7%-15% of their average values. This indicates that daily film calibrations are essential when EDR2 film is used to obtain absolute dose results. An elemental analysis of EDR2 film revealed that it contains 60% as much silver and 20% as much bromine as Kodak XV2 film. EDR2 film also has an unusual 1.69:1 silver:halide molar ratio, compared with the XV2 film's 1.02:1 ratio, which may affect its chemical reactions. To test EDR2's intermittency effect, the OD generated by a single 300 MU exposure was compared to the ODs generated by exposing the film 1 MU, 2 MU, and 4 MU at a time to a total of 300 MU. An ion chamber recorded the relative dose of all intermittency measurements to account for machine output variations. Using small MU bursts to expose the film resulted in delivery times of 4 to 14 minutes and lowered the film's OD by approximately 2% for both 6 and 18 MV beams. This effect may result in EDR2 film underestimating absolute doses for patient verifications that require long delivery times. After using a calibration to convert EDR2 film's OD to dose values, film measurements agreed within 2% relative difference and 2 mm criteria to ion chamber measurements for both sliding window and step-and-shoot fluence map verifications. Calibrated film results agreed with ion chamber measurements to within 5 % /2 mm criteria for transverse-plane full-plan verifications, but were consistently low. When properly calibrated, EDR2 film can be an adequate two-dimensional dosimeter for IMRT verifications, although it may underestimate doses in regions with long exposure times.

  1. Implementation of an anonymisation tool for clinical trials using a clinical trial processor integrated with an existing trial patient data information system.

    PubMed

    Aryanto, Kadek Y E; Broekema, André; Oudkerk, Matthijs; van Ooijen, Peter M A

    2012-01-01

    To present an adapted Clinical Trial Processor (CTP) test set-up for receiving, anonymising and saving Digital Imaging and Communications in Medicine (DICOM) data using external input from the original database of an existing clinical study information system to guide the anonymisation process. Two methods are presented for an adapted CTP test set-up. In the first method, images are pushed from the Picture Archiving and Communication System (PACS) using the DICOM protocol through a local network. In the second method, images are transferred through the internet using the HTTPS protocol. In total 25,000 images from 50 patients were moved from the PACS, anonymised and stored within roughly 2 h using the first method. In the second method, an average of 10 images per minute were transferred and processed over a residential connection. In both methods, no duplicated images were stored when previous images were retransferred. The anonymised images are stored in appropriate directories. The CTP can transfer and process DICOM images correctly in a very easy set-up providing a fast, secure and stable environment. The adapted CTP allows easy integration into an environment in which patient data are already included in an existing information system.

  2. Cone-Beam Computed Tomography Internal Motion Tracking Should Be Used to Validate 4-Dimensional Computed Tomography for Abdominal Radiation Therapy Patients

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Rankine, Leith; Wan, Hanlin; Parikh, Parag

    Purpose: To demonstrate that fiducial tracking during pretreatment Cone-Beam CT (CBCT) can accurately measure tumor motion and that this method should be used to validate 4-dimensional CT (4DCT) margins before each treatment fraction. Methods and Materials: For 31 patients with abdominal tumors and implanted fiducial markers, tumor motion was measured daily with CBCT and fluoroscopy for 202 treatment fractions. Fiducial tracking and maximum-likelihood algorithms extracted 3-dimensional fiducial trajectories from CBCT projections. The daily internal margin (IM) (ie, range of fiducial motion) was calculated for CBCT and fluoroscopy as the 5th-95th percentiles of displacement in each cardinal direction. The planning IMmore » from simulation 4DCT (IM{sub 4DCT}) was considered adequate when within ±1.2 mm (anterior–posterior, left–right) and ±3 mm (superior–inferior) of the daily measured IM. We validated CBCT fiducial tracking as an accurate predictive measure of intrafraction motion by comparing the daily measured IM{sub CBCT} with the daily IM measured by pretreatment fluoroscopy (IM{sub pre-fluoro}); these were compared with pre- and posttreatment fluoroscopy (IM{sub fluoro}) to identify those patients who could benefit from imaging during treatment. Results: Four-dimensional CT could not accurately predict intrafractional tumor motion for ≥80% of fractions in 94% (IM{sub CBCT}), 97% (IM{sub pre-fluoro}), and 100% (IM{sub fluoro}) of patients. The IM{sub CBCT} was significantly closer to IM{sub pre-fluoro} than IM{sub 4DCT} (P<.01). For patients with median treatment time t < 7.5 minutes, IM{sub CBCT} was in agreement with IM{sub fluoro} for 93% of fractions (superior–inferior), compared with 63% for the t > 7.5 minutes group, demonstrating the need for patient-specific intratreatment imaging. Conclusions: Tumor motion determined from 4DCT simulation does not accurately predict the daily motion observed on CBCT or fluoroscopy. Cone-beam CT could replace fluoroscopy for pretreatment verification of simulation IM{sub 4DCT}, reducing patient setup time and imaging dose. Patients with treatment time t > 7.5 minutes could benefit from the addition of intratreatment imaging.« less

  3. SU-E-T-632: A Dosimetric Comparison of the 3D-CRT Planning of Chest Wall in Post-Mastectomy Breast Cancer Patients, with and Without Breast Board Setup

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Muzaffar, Ambreen; Masood, Asif; Ullah, Haseeb

    2014-06-15

    Purpose: Breast boards are used in breast radiation which increases normal lung and heart doses, when supraclavicular field is included. Therefore, in this study through dose volume histogram (DVHs), lung and heart doses comparison was done between two different setups i.e. with and without breast board, for the treatment of left chest wall and supraclavicular fossa in postmastectomy left breast cancer. Methods: In this study, CT-Simulation scans of ten breast cancer patients were done with and without breast board, at Shifa International Hospitals Islamabad, to investigate the differences between the two different setups of the irradiation of left chest wallmore » in terms of lung and heart doses. For immobilization, support under the neck, shoulders and arms was used. Precise PLAN 2.15 treatment planning system (TPS) was used for 3D-CRT planning. The total prescribed dose for both the plans was 5000 cGy/25 fractions. The chest wall was treated with a pair of tangential photon fields and the upper supraclavicular nodal regions were treated with an anterior photon field. A mono-isocentric technique was used to match the tangential fields with the anterior field at the isocentre. The dose volume histogram was used to compare the doses of heart and ipsilateral lung. Results: Both the plans of each patient were generated and compared. DVH results showed that for the same PTV dose coverage, plans without breast board resulted in a reduction of lung and heart doses compared with the plans with breast board. There was significant reductions in V20, V<25 and mean doses for lung and V<9 and mean doses for heart. Conclusion: In comparison of both the plans, setup without breast board significantly reduced the dose-volume of the ipsilateral lung and heart in left chest wall patients. Waived registration request has been submitted.« less

  4. SU-E-T-560: Inter- and Intra-Fraction Variations in Esophageal Dose for Lung Cancer Patients, and the Impact of Setup Technique and Treatment Modality.

    PubMed

    Carroll, M; Cheung, J; Zhang, L; Court, L

    2012-06-01

    To understand the dose-response of the esophagus in photon and proton therapy, it is important to appreciate the variations in delivered dose caused by inter- and intra-fraction motion. Four lung cancer patients were identified who had experienced grade 3 esophagitis during their treatment, and for whom their esophagus was close, but not encompassed by, the treatment volume. Each patient had been treated with proton therapy using 35-37 2Gy fractions, and had received weekly 4DCT imaging. IMRT plans were also created using the same treatment planning constraints. In-house image registration software was used to deform the esophagus contour from the treatment plan to each phase of the 4DCT for each weekly image set. Daily setup using both bony and soft tissue (GTV) registration was simulated, and the treatment dose calculated for each CT image. Changes to the esophagus DVH relative to the treatment plan were quantified in terms of the relative volume of the esophagus receiving 45, 55, and 65Gy (V45, V55 and V65). For all combinations of treatment modality (photon, proton) and setup method (bony, GTV), intra-fraction motion resulted in a range of V45, V55 and V65 from 3.6 to 5.5%. Inter-fraction motion comparing daily exhale or inhale phases showed the range of V45, V55 and V65 from 8.5 to 18.6% (exhale) and 9.8 to 16.3% (inhale). Inter-fractional motion resulted in larger variations in dose delivered to the esophagus than intra-fractional motion. The inter-fraction range for V45, V55 and V65 varied by around 10% between patients. The treatment modality (photon, proton) and setup technique (bony, GTV) had minimal impact on the results. © 2012 American Association of Physicists in Medicine.

  5. SU-F-T-509: Investigation into the Impact of the Linear Accelerator Treatment Table On Skin Dose to Prone Breast Patients

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Pedersen, K; Irwin, J; Sansourekidou, P

    Purpose: To investigate the impact of the treatment table on skin dose for prone breast patients for which the breast contacts the table and to develop a method to decrease skin dose. Methods: We used 12cm stack of 15cmx15cm solid water slabs to imitate breast. Calibrated EBT3 radiochromic film was affixed to the bottom of the phantom. Treatments for 32 patients were analyzed to determine typical prone breast beam parameters. Based on the analysis, a field size and a range of gantry angles were chosen for the test beams. Three experimental setups were used. The first represented the patient setupmore » currently used in our clinics with the phantom directly on the table. The second was the skin sparing setup, with a 1.5cm Styrofoam slab between the phantom and the table. The third used a 7.5cm Styrofoam slab to examine the extent of skin sparing potential. The calibration curve was applied to each film to determine dose. Percent difference in dose between the current and skin sparing setups was calculated for each gantry angle and gantry angle pair. Results: Data showed that beams entering through the table showed a skin dose decrease ranging from 13%–30% with the addition of 7.5cm Styrofoam, while beams exiting through the table showed no significant difference. The addition of 1.5cm Styrofoam resulted in differences ranging from 0.5%–13% with the skin sparing setup. Conclusion: The results demonstrate that skin in contact with the table receives increased dose from beams entering through the table. By creating separation between the breast and the table with Styrofoam the skin dose can be lowered, but 1.5 cm did not fully mitigate the effect. Further investigation will be performed to identify a clinically practical thickness that maximizes this mitigation.« less

  6. TU-H-CAMPUS-JeP1-02: Fully Automatic Verification of Automatically Contoured Normal Tissues in the Head and Neck

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    McCarroll, R; UT Health Science Center, Graduate School of Biomedical Sciences, Houston, TX; Beadle, B

    Purpose: To investigate and validate the use of an independent deformable-based contouring algorithm for automatic verification of auto-contoured structures in the head and neck towards fully automated treatment planning. Methods: Two independent automatic contouring algorithms [(1) Eclipse’s Smart Segmentation followed by pixel-wise majority voting, (2) an in-house multi-atlas based method] were used to create contours of 6 normal structures of 10 head-and-neck patients. After rating by a radiation oncologist, the higher performing algorithm was selected as the primary contouring method, the other used for automatic verification of the primary. To determine the ability of the verification algorithm to detect incorrectmore » contours, contours from the primary method were shifted from 0.5 to 2cm. Using a logit model the structure-specific minimum detectable shift was identified. The models were then applied to a set of twenty different patients and the sensitivity and specificity of the models verified. Results: Per physician rating, the multi-atlas method (4.8/5 point scale, with 3 rated as generally acceptable for planning purposes) was selected as primary and the Eclipse-based method (3.5/5) for verification. Mean distance to agreement and true positive rate were selected as covariates in an optimized logit model. These models, when applied to a group of twenty different patients, indicated that shifts could be detected at 0.5cm (brain), 0.75cm (mandible, cord), 1cm (brainstem, cochlea), or 1.25cm (parotid), with sensitivity and specificity greater than 0.95. If sensitivity and specificity constraints are reduced to 0.9, detectable shifts of mandible and brainstem were reduced by 0.25cm. These shifts represent additional safety margins which might be considered if auto-contours are used for automatic treatment planning without physician review. Conclusion: Automatically contoured structures can be automatically verified. This fully automated process could be used to flag auto-contours for special review or used with safety margins in a fully automatic treatment planning system.« less

  7. A Quality Assurance Method that Utilizes 3D Dosimetry and Facilitates Clinical Interpretation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Oldham, Mark, E-mail: mark.oldham@duke.edu; Thomas, Andrew; O'Daniel, Jennifer

    2012-10-01

    Purpose: To demonstrate a new three-dimensional (3D) quality assurance (QA) method that provides comprehensive dosimetry verification and facilitates evaluation of the clinical significance of QA data acquired in a phantom. Also to apply the method to investigate the dosimetric efficacy of base-of-skull (BOS) intensity-modulated radiotherapy (IMRT) treatment. Methods and Materials: Two types of IMRT QA verification plans were created for 6 patients who received BOS IMRT. The first plan enabled conventional 2D planar IMRT QA using the Varian portal dosimetry system. The second plan enabled 3D verification using an anthropomorphic head phantom. In the latter, the 3D dose distribution wasmore » measured using the DLOS/Presage dosimetry system (DLOS = Duke Large-field-of-view Optical-CT System, Presage Heuris Pharma, Skillman, NJ), which yielded isotropic 2-mm data throughout the treated volume. In a novel step, measured 3D dose distributions were transformed back to the patient's CT to enable calculation of dose-volume histograms (DVH) and dose overlays. Measured and planned patient DVHs were compared to investigate clinical significance. Results: Close agreement between measured and calculated dose distributions was observed for all 6 cases. For gamma criteria of 3%, 2 mm, the mean passing rate for portal dosimetry was 96.8% (range, 92.0%-98.9%), compared to 94.9% (range, 90.1%-98.9%) for 3D. There was no clear correlation between 2D and 3D passing rates. Planned and measured dose distributions were evaluated on the patient's anatomy, using DVH and dose overlays. Minor deviations were detected, and the clinical significance of these are presented and discussed. Conclusions: Two advantages accrue to the methods presented here. First, treatment accuracy is evaluated throughout the whole treated volume, yielding comprehensive verification. Second, the clinical significance of any deviations can be assessed through the generation of DVH curves and dose overlays on the patient's anatomy. The latter step represents an important development that advances the clinical relevance of complex treatment QA.« less

  8. Clinical benefits of new immobilization system for hypofractionated radiotherapy of intrahepatic hepatocellular carcinoma by helical tomotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hu, Yong; Zhou, Yong-Kang; Chen, Yi-Xing

    Objective: A comprehensive clinical evaluation was conducted, assessing the Body Pro-Lok immobilization and positioning system to facilitate hypofractionated radiotherapy of intrahepatic hepatocellular carcinoma (HCC), using helical tomotherapy to improve treatment precision. Methods: Clinical applications of the Body Pro-Lok system were investigated (as above) in terms of interfractional and intrafractional setup errors and compressive abdominal breath control. To assess interfractional setup errors, a total of 42 patients who were given 5 to 20 fractions of helical tomotherapy for intrahepatic HCC were analyzed. Overall, 15 patients were immobilized using simple vacuum cushion (group A), and the Body Pro-Lok system was used inmore » 27 patients (group B), performing megavoltage computed tomography (MVCT) scans 196 times and 435 times, respectively. Pretreatment MVCT scans were registered to the planning kilovoltage computed tomography (KVCT) for error determination, and group comparisons were made. To establish intrafractional setup errors, 17 patients with intrahepatic HCC were selected at random for immobilization by Body Pro-Lok system, undergoing MVCT scans after helical tomotherapy every week. A total of 46 MVCT re-scans were analyzed for this purpose. In researching breath control, 12 patients, randomly selected, were immobilized by Body Pro-Lok system and subjected to 2-phase 4-dimensional CT (4DCT) scans, with compressive abdominal control or in freely breathing states, respectively. Respiratory-induced liver motion was then compared. Results: Mean interfractional setup errors were as follows: (1) group A: X, 2.97 ± 2.47 mm; Y, 4.85 ± 4.04 mm; and Z, 3.77 ± 3.21 mm; pitch, 0.66 ± 0.62°; roll, 1.09 ± 1.06°; and yaw, 0.85 ± 0.82°; and (2) group B: X, 2.23 ± 1.79 mm; Y, 4.10 ± 3.36 mm; and Z, 1.67 ± 1.91 mm; pitch, 0.45 ± 0.38°; roll, 0.77 ± 0.63°; and yaw, 0.52 ± 0.49°. Between-group differences were statistically significant in 6 directions (p < 0.05). Mean intrafractional setup errors with use of the Body Pro-Lok system were as follows: X, 0.41 ± 0.46 mm; Y, 0.86 ± 0.80 mm; Z, 0.33 ± 0.44 mm; and roll, 0.12 ± 0.19°. Mean liver-induced respiratory motion determinations were as follows: (1) abdominal compression: X, 2.33 ± 1.22 mm; Y, 5.11 ± 2.05 mm; Z, 2.13 ± 1.05 mm; and 3D vector, 6.22 ± 1.94 mm; and (2) free breathing: X, 3.48 ± 1.14 mm; Y, 9.83 ± 3.00 mm; Z, 3.38 ± 1.59 mm; and 3D vector, 11.07 ± 3.16 mm. Between-group differences were statistically different in 4 directions (p < 0.05). Conclusions: The Body Pro-Lok system is capable of improving interfractional and intrafractional setup accuracy and minimizing tumor movement owing to respirations in patients with intrahepatic HCC during hypofractionated helical tomotherapy.« less

  9. New developments in EPID-based 3D dosimetry in The Netherlands Cancer Institute

    NASA Astrophysics Data System (ADS)

    Mijnheer, B.; Rozendaal, R.; Olaciregui-Ruiz, I.; González, P.; van Oers, R.; Mans, A.

    2017-05-01

    EPID-based offline 3D in vivo dosimetry is performed routinely in The Netherlands Cancer Institute for almost all RT treatments. The 3D dose distribution is reconstructed using the EPID primary dose in combination with a back-projection algorithm and compared with the planned dose distribution. Recently the method was adapted for real-time dose verification, performing 3D dose verification in less than 300 ms, which is faster than the current portal frame acquisition rate. In this way a possibility is created for halting the linac in case of large delivery errors. Furthermore, a new method for pre-treatment QA was developed in which the EPID primary dose behind a phantom or patient is predicted using the CT data of that phantom or patient in combination with in-air EPID measurements. This virtual EPID primary transit dose is then used to reconstruct the 3D dose distribution within the phantom or patient geometry using the same dose engine as applied offline. In order to assess the relevance of our clinically applied alert criteria, we investigated the sensitivity of our EPID-based 3D dose verification system to detect delivery errors in VMAT treatments. This was done through simulation by modifying patient treatment plans, as well as experimentally by performing EPID measurements during the irradiation of an Alderson phantom, both after deliberately introducing errors during VMAT delivery. In this presentation these new developments will be elucidated.

  10. Teaching Cancer Patients the Value of Correct Positioning During Radiotherapy Using Visual Aids and Practical Exercises.

    PubMed

    Hansen, Helle; Nielsen, Berit Kjærside; Boejen, Annette; Vestergaard, Anne

    2018-06-01

    The aim of this study was to investigate if teaching patients about positioning before radiotherapy treatment would (a) reduce the residual rotational set-up errors, (b) reduce the number of repositionings and (c) improve patients' sense of control by increasing self-efficacy and reducing distress. Patients were randomized to either standard care (control group) or standard care and a teaching session combining visual aids and practical exercises (intervention group). Daily images from the treatment sessions were evaluated off-line. Both groups filled in a questionnaire before and at the end of the treatment course on various aspects of cooperation with the staff regarding positioning. Comparisons of residual rotational set-up errors showed an improvement in the intervention group compared to the control group. No significant differences were found in number of repositionings, self-efficacy or distress. Results show that it is possible to teach patients about positioning and thereby improve precision in positioning. Teaching patients about positioning did not seem to affect self-efficacy or distress scores at baseline and at the end of the treatment course.

  11. The application of a low-cost 3D depth camera for patient set-up and respiratory motion management in radiotherapy

    NASA Astrophysics Data System (ADS)

    Tahavori, Fatemeh

    Respiratory motion induces uncertainty in External Beam Radiotherapy (EBRT), which can result in sub-optimal dose delivery to the target tissue and unwanted dose to normal tissue. The conventional approach to managing patient respiratory motion for EBRT within the area of abdominal-thoracic cancer is through the use of internal radiological imaging methods (e.g. Megavoltage imaging or Cone-Beam Computed Tomography) or via surrogate estimates of tumour position using external markers placed on the patient chest. This latter method uses tracking with video-based techniques, and relies on an assumed correlation or mathematical model, between the external surrogate signal and the internal target position. The marker's trajectory can be used in both respiratory gating techniques and real-time tracking methods. Internal radiological imaging methods bring with them limited temporal resolution, and additional radiation burden, which can be addressed by external marker-based methods that carry no such issues. Moreover, by including multiple external markers and placing them closer to the internal target organs, the effciency of correlation algorithms can be increased. However, the quality of such external monitoring methods is underpinned by the performance of the associated correlation model. Therefore, several new approaches to correlation modelling have been developed as part of this thesis and compared using publicly-available datasets. Highly competitive results have been obtained when compared against state-of-the-art methods. Marker-based methods also have the disadvantages of requiring manual set-up time for marker placement and patient positioning and potential issues with reproducibility of marker placement. This motivates the investigation of non-contact marker-free methods for use in EBRT, which is the main topic of this thesis. The Microsoft Kinect is used as an example of a low-cost consumer grade 3D depth camera for capturing and analysing external respiratory motion. This thesis makes the first presentation of detailed studies of external respiratory motion captured using such low-cost technology and demonstrates its potential in a healthcare environment. Firstly, the fundamental performance of a range of Microsoft Kinect sensors is assessed for use in radiotherapy (and potentially other healthcare applications), in terms of static and dynamic performance using both phantoms and volunteers. Then external respiratory motion is captured using the above technology from a group of 32 healthy volunteers and Principal Component Analysis (PCA) is applied to a region of interest encompassing the complete anterior surface to demonstrate breathing style. This work demonstrates that this surface motion can be compactly described by the first two PCA eigenvectors. The reproducibility of subject-specific EBRT set-up using conventional laser-based alignment and marker-based Deep Inspiration Breath Hold (DIBH) methods are also studied using the Microsoft Kinect sensor. A cohort of five healthy female volunteers is repeatedly set-up for left-sided breast cancer EBRT and multiple DIBH episodes captured over five separate sessions representing multiple fractionated radiotherapy treatment sessions, but without dose delivery. This provided an independent assessment that subjects were set-up and generally achieved variations within currently accepted margins of clinical practice. Moreover, this work demonstrated the potential role of consumer-grade 3D depth camera technology as a possible replacement for marker based set-up and DIBH management procedures. This brings with it the additional benefits of low cost, and potential through-put benefits, as patient set-up could ultimately be fully automated with this technology, and DIBH could be independently monitored without requiring preparatory manual intervention.

  12. On-Board Imaging Validation of Optically Guided Stereotactic Radiosurgery Positioning System for Conventionally Fractionated Radiotherapy for Paranasal Sinus and Skull Base Cancer

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Maxim, Peter G.; Loo, Billy W.; Murphy, James D.

    2011-11-15

    Purpose: To evaluate the positioning accuracy of an optical positioning system for stereotactic radiosurgery in a pilot experience of optically guided, conventionally fractionated, radiotherapy for paranasal sinus and skull base tumors. Methods and Materials: Before each daily radiotherapy session, the positioning of 28 patients was set up using an optical positioning system. After this initial setup, the patients underwent standard on-board imaging that included daily orthogonal kilovoltage images and weekly cone beam computed tomography scans. Daily translational shifts were made after comparing the on-board images with the treatment planning computed tomography scans. These daily translational shifts represented the daily positionalmore » error in the optical tracking system and were recorded during the treatment course. For 13 patients treated with smaller fields, a three-degree of freedom (3DOF) head positioner was used for more accurate setup. Results: The mean positional error for the optically guided system in patients with and without the 3DOF head positioner was 1.4 {+-} 1.1 mm and 3.9 {+-} 1.6 mm, respectively (p <.0001). The mean positional error drifted 0.11 mm/wk upward during the treatment course for patients using the 3DOF head positioner (p = .057). No positional drift was observed in the patients without the 3DOF head positioner. Conclusion: Our initial clinical experience with optically guided head-and-neck fractionated radiotherapy was promising and demonstrated clinical feasibility. The optically guided setup was especially useful when used in conjunction with the 3DOF head positioner and when it was recalibrated to the shifts using the weekly portal images.« less

  13. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kim, T; Cho, M; Kang, S

    Purpose: To improve the setup accuracy of thermoplastic mask, we developed a new monitoring method based on force sensing technology and evaluated its feasibility. Methods: The thermoplastic mask setup monitoring system consists of a force sensing resistor sensor unit, a signal transport device, a control PC and an in-house software. The system is designed to monitor pressure variation between the mask and patient in real time. It also provides a warning to the user when there is a possibility of movement. A preliminary study was performed to evaluate the reliability of the sensor unit and developed monitoring system with amore » head phantom. Then, a simulation study with volunteers was conducted to evaluate the feasibility of the monitoring system. Note that the sensor unit can have multiple end-sensors and every end-sensor was confirmed to be within 2% reliability in pressure reading through a screening test. Results: To evaluate the reproducibility of the proposed monitoring system in practice, we simulated a mask setup with the head phantom. FRS sensors were attached on the face of the head phantom and pressure was monitored. For 3 repeated mask setups on the phantom, the variation of the pressure was less than 3% (only 1% larger than 2% potential uncertainty confirmed in the screening test). In the volunteer study, we intended to verify that the system could detect patient movements within the mask. Thus, volunteers were asked to turn their head or lift their chin. The system was able to detect movements effectively, confirming the clinical feasibility of the monitoring system developed. Conclusion: Through the proposed setup monitoring method, it is possible to monitor patient motion inside a mask in real time, which has never been possible with most commonly used systems using non-radiographic technology such as infrared camera system and surface imaging system. This work was supported by the Radiation Technology R&D program (No. 2013M2A2A7043498) and the Mid-career Researcher Program (2014R1A2A1A10050270) through the National Research Foundation of Korea funded by the Ministry of Science, ICT&Future Planning.« less

  14. SU-E-J-150: Impact of Intrafractional Prostate Motion On the Accuracy and Efficiency of Prostate SBRT Delivery: A Retrospective Analysis of Prostate Tracking Log Files

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Xiang, H; Hirsch, A; Willins, J

    2014-06-01

    Purpose: To measure intrafractional prostate motion by time-based stereotactic x-ray imaging and investigate the impact on the accuracy and efficiency of prostate SBRT delivery. Methods: Prostate tracking log files with 1,892 x-ray image registrations from 18 SBRT fractions for 6 patients were retrospectively analyzed. Patient setup and beam delivery sessions were reviewed to identify extended periods of large prostate motion that caused delays in setup or interruptions in beam delivery. The 6D prostate motions were compared to the clinically used PTV margin of 3–5 mm (3 mm posterior, 5 mm all other directions), a hypothetical PTV margin of 2–3 mmmore » (2 mm posterior, 3 mm all other directions), and the rotation correction limits (roll ±2°, pitch ±5° and yaw ±3°) of CyberKnife to quantify beam delivery accuracy. Results: Significant incidents of treatment start delay and beam delivery interruption were observed, mostly related to large pitch rotations of ≥±5°. Optimal setup time of 5–15 minutes was recorded in 61% of the fractions, and optimal beam delivery time of 30–40 minutes in 67% of the fractions. At a default imaging interval of 15 seconds, the percentage of prostate motion beyond PTV margin of 3–5 mm varied among patients, with a mean at 12.8% (range 0.0%–31.1%); and the percentage beyond PTV margin of 2–3 mm was at a mean of 36.0% (range 3.3%–83.1%). These timely detected offsets were all corrected real-time by the robotic manipulator or by operator intervention at the time of treatment interruptions. Conclusion: The durations of patient setup and beam delivery were directly affected by the occurrence of large prostate motion. Frequent imaging of down to 15 second interval is necessary for certain patients. Techniques for reducing prostate motion, such as using endorectal balloon, can be considered to assure consistently higher accuracy and efficiency of prostate SBRT delivery.« less

  15. Effects of computerized prescriber order entry on pharmacy order-processing time.

    PubMed

    Wietholter, Jon; Sitterson, Susan; Allison, Steven

    2009-08-01

    The effect of computerized prescriber order entry (CPOE) on the efficiency of medication-order-processing time was evaluated. This study was conducted at a 761-bed, tertiary care hospital. A total of 2988 medication orders were collected and analyzed before (n = 1488) and after CPOE implementation (n = 1500). Data analyzed included the time the prescriber ordered the medication, the time the pharmacy received the order, and the time the order was completed by a pharmacist. The mean order-processing time before CPOE implementation was 115 minutes from prescriber composition to pharmacist verification. After CPOE implementation, the mean order-processing time was reduced to 3 minutes (p < 0.0001). The time that an order was received by the pharmacy to the time it was verified by a pharmacist was reduced from 31 minutes before CPOE implementation to 3 minutes after CPOE implementation (p < 0.0001). The implementation of CPOE reduced the order-processing time (from order composition to verification) by 97%. Additionally, pharmacy-specific order-processing time (from order receipt in the pharmacy to pharmacist verification) was reduced by 90%. This reduction in order-processing time improves patient care by shortening the interval between physician prescribing and medication availability and may allow pharmacists to explore opportunities for enhanced clinical activities that will further positively impact patient care. CPOE implementation reduced the mean pharmacy order-processing time from composition to verification by 97%. After CPOE implementation, a new medication order was verified as appropriate by a pharmacist in three minutes, on average.

  16. Development of a multi-knife-edge slit collimator for prompt gamma ray imaging during proton beam cancer therapy

    NASA Astrophysics Data System (ADS)

    Ready, John Francis, III

    Proton beam usage to treat cancer has recently experienced rapid growth, as it offers the ability to target dose delivery in a patient more precisely than traditional x-ray treatment methods. Protons stop within the patient, delivering the maximum dose at the end of their track--a phenomenon described as the Bragg peak. However, because a large dose is delivered to a small volume, proton therapy is very sensitive to errors in patient setup and treatment planning calculations. Additionally, because all primary beam particles stop in the patient, there is no direct information available to verify dose delivery. These factors contribute to the range uncertainty in proton therapy, which ultimately hinders its clinical usefulness. A reliable method of proton range verification would allow the clinician to fully utilize the precise dose delivery of the Bragg peak. Several methods to verify proton range detect secondary emissions, especially prompt gamma ray (PG) emissions. However, detection of PGs is challenging due to their high energy (2-10 MeV) and low attenuation coefficients, which limit PG interactions in materials. Therefore, detection and collimation methods must be specifically designed for prompt gamma ray imaging (PGI) applications. In addition, production of PGs relies on delivering a dose of radiation to the patient. Ideally, verification of the Bragg peak location exposes patients to a minimal dose, thus limiting the PG counts available to the imaging system. An additional challenge for PGI is the lack of accurate simulation models, which limit the study of PG production characteristics and the relationship between PG distribution and dose delivery. Specific limitations include incorrect modeling of the reaction cross sections, gamma emission yields, and angular distribution of emission for specific photon energies. While simulations can still be valuable assets in designing a system to detect and image PGs, until new models are developed and incorporated into Monte Carlo simulation packages, simulations cannot be used to study the production and location of PG emissions during proton therapy. This work presents a novel system to image PGs emitted during proton therapy to verify proton beam range. The imaging system consists of a multi-slit collimator paired with a position-sensitive LSO scintillation detector. This innovative design is the first collimated imaging system to implement two-dimensional (2-D) imaging for PG proton beam range verification, while also providing a larger field of view than compared to single-slit collimator systems. Other, uncollimated imaging systems have been explored for PGI applications, such as Compton cameras. However, Compton camera designs are severely limited by counting rate capabilities. A recent Compton camera study reported count rate capability of about 5 kHz. However, at a typical clinical beam current of 1.0 nA, the estimated PG emission rate would be 6 x 108 per second. After accounting for distance to the detector and interaction efficiencies, the detection system will still be overwhelmed with counts in the MHz range, causing false coincidences and hindering the operation of the imaging system. Initial measurements using 50 MeV protons demonstrated the ability of our system to reconstruct 2-D PG distributions at clinical beam currents. A Bragg peak localization precision of 1 mm (2sigma) was achieved with delivery of (1.7 +/- 0.8) x 108 protons into a PMMA target, suggesting the ability of the system to detect relative shifts in proton range while delivering fewer protons than used in a typical treatment fraction. This is key, as the ideal system allows the clinician to verify proton range when delivering only a small portion of the prescribed dose, preventing the mistreatment of the patient. Additionally, the absolute position of the Bragg peak was identified to within 1.6 mm (2sigma) with 5.6 x 1010 protons delivered. These promising results warrant further investigation and system optimization for clinical implementation. While further measurements at clinical beam energy levels will be required to verify system performance, these preliminary results provide evidence that 2-D image reconstruction, with 1-2 mm accuracy, is possible with this design. Implementing such a system in the clinical setting would greatly improve proton therapy cancer treatment outcomes.

  17. SU-E-T-435: Development and Commissioning of a Complete System for In-Vivo Dosimetry and Range Verification in Proton Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Samuel, D; Testa, M; Park, Y

    Purpose: In-vivo dose and beam range verification in proton therapy could play significant roles in proton treatment validation and improvements. Invivo beam range verification, in particular, could enable new treatment techniques one of which, for example, could be the use of anterior fields for prostate treatment instead of opposed lateral fields as in current practice. We have developed and commissioned an integrated system with hardware, software and workflow protocols, to provide a complete solution, simultaneously for both in-vivo dosimetry and range verification for proton therapy. Methods: The system uses a matrix of diodes, up to 12 in total, but separablemore » into three groups for flexibility in application. A special amplifier was developed to capture extremely small signals from very low proton beam current. The software was developed within iMagX, a general platform for image processing in radiation therapy applications. The range determination exploits the inherent relationship between the internal range modulation clock of the proton therapy system and the radiological depth at the point of measurement. The commissioning of the system, for in-vivo dosimetry and for range verification was separately conducted using anthropomorphic phantom. EBT films and TLDs were used for dose comparisons and range scan of the beam distal fall-off was used as ground truth for range verification. Results: For in-vivo dose measurement, the results were in agreement with TLD and EBT films and were within 3% from treatment planning calculations. For range verification, a precision of 0.5mm is achieved in homogeneous phantoms, and a precision of 2mm for anthropomorphic pelvic phantom, except at points with significant range mixing. Conclusion: We completed the commissioning of our system for in-vivo dosimetry and range verification in proton therapy. The results suggest that the system is ready for clinical trials on patient.« less

  18. Quantitative assessment of the physical potential of proton beam range verification with PET/CT.

    PubMed

    Knopf, A; Parodi, K; Paganetti, H; Cascio, E; Bonab, A; Bortfeld, T

    2008-08-07

    A recent clinical pilot study demonstrated the feasibility of offline PET/CT range verification for proton therapy treatments. In vivo PET measurements are challenged by blood perfusion, variations of tissue compositions, patient motion and image co-registration uncertainties. Besides these biological and treatment specific factors, the accuracy of the method is constrained by the underlying physical processes. This phantom study distinguishes physical factors from other factors, assessing the reproducibility, consistency and sensitivity of the PET/CT range verification method. A spread-out Bragg-peak (SOBP) proton field was delivered to a phantom consisting of poly-methyl methacrylate (PMMA), lung and bone equivalent material slabs. PET data were acquired in listmode at a commercial PET/CT scanner available within 10 min walking distance from the proton therapy unit. The measured PET activity distributions were compared to simulations of the PET signal based on Geant4 and FLUKA Monte Carlo (MC) codes. To test the reproducibility of the measured PET signal, data from two independent measurements at the same geometrical position in the phantom were compared. Furthermore, activation depth profiles within identical material arrangements but at different positions within the irradiation field were compared to test the consistency of the measured PET signal. Finally, activation depth profiles through air/lung, air/bone and lung/bone interfaces parallel as well as at 6 degrees to the beam direction were studied to investigate the sensitivity of the PET/CT range verification method. The reproducibility and the consistency of the measured PET signal were found to be of the same order of magnitude. They determine the physical accuracy of the PET measurement to be about 1 mm. However, range discrepancies up to 2.6 mm between two measurements and range variations up to 2.6 mm within one measurement were found at the beam edge and at the edge of the field of view (FOV) of the PET scanner. PET/CT range verification was found to be able to detect small range modifications in the presence of complex tissue inhomogeneities. This study indicates the physical potential of the PET/CT verification method to detect the full-range characteristic of the delivered dose in the patient.

  19. Quantitative assessment of the physical potential of proton beam range verification with PET/CT

    NASA Astrophysics Data System (ADS)

    Knopf, A.; Parodi, K.; Paganetti, H.; Cascio, E.; Bonab, A.; Bortfeld, T.

    2008-08-01

    A recent clinical pilot study demonstrated the feasibility of offline PET/CT range verification for proton therapy treatments. In vivo PET measurements are challenged by blood perfusion, variations of tissue compositions, patient motion and image co-registration uncertainties. Besides these biological and treatment specific factors, the accuracy of the method is constrained by the underlying physical processes. This phantom study distinguishes physical factors from other factors, assessing the reproducibility, consistency and sensitivity of the PET/CT range verification method. A spread-out Bragg-peak (SOBP) proton field was delivered to a phantom consisting of poly-methyl methacrylate (PMMA), lung and bone equivalent material slabs. PET data were acquired in listmode at a commercial PET/CT scanner available within 10 min walking distance from the proton therapy unit. The measured PET activity distributions were compared to simulations of the PET signal based on Geant4 and FLUKA Monte Carlo (MC) codes. To test the reproducibility of the measured PET signal, data from two independent measurements at the same geometrical position in the phantom were compared. Furthermore, activation depth profiles within identical material arrangements but at different positions within the irradiation field were compared to test the consistency of the measured PET signal. Finally, activation depth profiles through air/lung, air/bone and lung/bone interfaces parallel as well as at 6° to the beam direction were studied to investigate the sensitivity of the PET/CT range verification method. The reproducibility and the consistency of the measured PET signal were found to be of the same order of magnitude. They determine the physical accuracy of the PET measurement to be about 1 mm. However, range discrepancies up to 2.6 mm between two measurements and range variations up to 2.6 mm within one measurement were found at the beam edge and at the edge of the field of view (FOV) of the PET scanner. PET/CT range verification was found to be able to detect small range modifications in the presence of complex tissue inhomogeneities. This study indicates the physical potential of the PET/CT verification method to detect the full-range characteristic of the delivered dose in the patient.

  20. Dosimetry investigation of MOSFET for clinical IMRT dose verification.

    PubMed

    Deshpande, Sudesh; Kumar, Rajesh; Ghadi, Yogesh; Neharu, R M; Kannan, V

    2013-06-01

    In IMRT, patient-specific dose verification is followed regularly at each centre. Simple and efficient dosimetry techniques play a very important role in routine clinical dosimetry QA. The MOSFET dosimeter offers several advantages over the conventional dosimeters such as its small detector size, immediate readout, immediate reuse, multiple point dose measurements. To use the MOSFET as routine clinical dosimetry system for pre-treatment dose verification in IMRT, a comprehensive set of experiments has been conducted, to investigate its linearity, reproducibility, dose rate effect and angular dependence for 6 MV x-ray beam. The MOSFETs shows a linear response with linearity coefficient of 0.992 for a dose range of 35 cGy to 427 cGy. The reproducibility of the MOSFET was measured by irradiating the MOSFET for ten consecutive irradiations in the dose range of 35 cGy to 427 cGy. The measured reproducibility of MOSFET was found to be within 4% up to 70 cGy and within 1.4% above 70 cGy. The dose rate effect on the MOSFET was investigated in the dose rate range 100 MU/min to 600 MU/min. The response of the MOSFET varies from -1.7% to 2.1%. The angular responses of the MOSFETs were measured at 10 degrees intervals from 90 to 270 degrees in an anticlockwise direction and normalized at gantry angle zero and it was found to be in the range of 0.98 ± 0.014 to 1.01 ± 0.014. The MOSFETs were calibrated in a phantom which was later used for IMRT verification. The measured calibration coefficients were found to be 1 mV/cGy and 2.995 mV/cGy in standard and high sensitivity mode respectively. The MOSFETs were used for pre-treatment dose verification in IMRT. Nine dosimeters were used for each patient to measure the dose in different plane. The average variation between calculated and measured dose at any location was within 3%. Dose verification using MOSFET and IMRT phantom was found to quick and efficient and well suited for a busy radiotherapy department.

  1. Dosimetric impact of daily setup variations during treatment of canine nasal tumors using intensity-modulated radiation therapy.

    PubMed

    Deveau, Michael A; Gutiérrez, Alonso N; Mackie, Thomas R; Tomé, Wolfgang A; Forrest, Lisa J

    2010-01-01

    Intensity-modulated radiation therapy (IMRT) can be employed to yield precise dose distributions that tightly conform to targets and reduce high doses to normal structures by generating steep dose gradients. Because of these sharp gradients, daily setup variations may have an adverse effect on clinical outcome such that an adjacent normal structure may be overdosed and/or the target may be underdosed. This study provides a detailed analysis of the impact of daily setup variations on optimized IMRT canine nasal tumor treatment plans when variations are not accounted for due to the lack of image guidance. Setup histories of ten patients with nasal tumors previously treated using helical tomotherapy were replanned retrospectively to study the impact of daily setup variations on IMRT dose distributions. Daily setup shifts were applied to IMRT plans on a fraction-by-fraction basis. Using mattress immobilization and laser alignment, mean setup error magnitude in any single dimension was at least 2.5 mm (0-10.0 mm). With inclusions of all three translational coordinates, mean composite offset vector was 5.9 +/- 3.3 mm. Due to variations, a loss of equivalent uniform dose for target volumes of up to 5.6% was noted which corresponded to a potential loss in tumor control probability of 39.5%. Overdosing of eyes and brain was noted by increases in mean normalized total dose and highest normalized dose given to 2% of the volume. Findings suggest that successful implementation of canine nasal IMRT requires daily image guidance to ensure accurate delivery of precise IMRT distributions when non-rigid immobilization techniques are utilized. Unrecognized geographical misses may result in tumor recurrence and/or radiation toxicities to the eyes and brain.

  2. DOSIMETRIC IMPACT OF DAILY SETUP VARIATIONS DURING TREATMENT OF CANINE NASAL TUMORS USING INTENSITY-MODULATED RADIATION THERAPY

    PubMed Central

    Deveau, Michael A.; Gutiérrez, Alonso N.; Mackie, Thomas R.; Tomé, Wolfgang A.; Forrest, Lisa J.

    2009-01-01

    Intensity-modulated radiation therapy (IMRT) can be employed to yield precise dose distributions that tightly conform to targets and reduce high doses to normal structures by generating steep dose gradients. Because of these sharp gradients, daily setup variations may have an adverse effect on clinical outcome such that an adjacent normal structure may be overdosed and/or the target may be underdosed. This study provides a detailed analysis of the impact of daily setup variations on optimized IMRT canine nasal tumor treatment plans when variations are not accounted for due to the lack of image guidance. Setup histories of ten patients with nasal tumors previously treated using helical tomotherapy were replanned retrospectively to study the impact of daily setup variations on IMRT dose distributions. Daily setup shifts were applied to IMRT plans on a fraction-by-fraction basis. Using mattress immobilization and laser alignment, mean setup error magnitude in any single dimension was at least 2.5mm (0-10.0mm). With inclusions of all three translational coordinates, mean composite offset vector was 5.9±3.3mm. Due to variations, a loss of equivalent uniform dose (EUD) for target volumes of up to 5.6% was noted which corresponded to a potential loss in TCP of 39.5%. Overdosing of eyes and brain was noted by increases in mean normalized total dose (NTDmean) and highest normalized dose given to 2% of the volume (NTD2%). Findings suggest that successful implementation of canine nasal IMRT requires daily image guidance to ensure accurate delivery of precise IMRT distributions when non-rigid immobilization techniques are utilized. Unrecognized geographical misses may result in tumor recurrence and/or radiation toxicities to the eyes and brain. PMID:20166402

  3. Patterns of intrafractional motion and uncertainties of treatment setup reference systems in accelerated partial breast irradiation for right- and left-sided breast cancer.

    PubMed

    Yue, Ning J; Goyal, Sharad; Kim, Leonard H; Khan, Atif; Haffty, Bruce G

    2014-01-01

    This study investigated the patterns of intrafractional motion and accuracy of treatment setup strategies in 3-dimensional conformal radiation therapy of accelerated partial breast irradiation (APBI) for right- and left-sided breast cancers. Sixteen right-sided and 17 left-sided breast cancer patients were enrolled in an institutional APBI trial in which gold fiducial markers were strategically sutured to the surgical cavity walls. Daily pre- and postradiation therapy kV imaging were performed and were matched to digitally reconstructed radiographs based on bony anatomy and fiducial markers, respectively, to determine the intrafractional motion. The positioning differences of the laser-tattoo and the bony anatomy-based setups with respect to the marker-based setup (benchmark) were determined to evaluate their accuracy. Statistical differences were found between the right- and left-sided APBI treatments in vector directions of intrafractional motion and treatment setup errors in the reference systems, but less in their overall magnitudes. The directional difference was more pronounced in the lateral direction. It was found that the intrafractional motion and setup reference systems tended to deviate in the right direction for the right-sided breast treatments and in the left direction for the left-sided breast treatments. It appears that the fiducial markers placed in the seroma cavity exhibit side dependent directional intrafractional motion, although additional data may be needed to further validate the conclusion. The bony anatomy-based treatment setup improves the accuracy over laser-tattoo. But it is inadequate to rely on bony anatomy to assess intrafractional target motion in both magnitude and direction. Copyright © 2014 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.

  4. A concept to transfer a therapeutic splint position into permanent occlusion with a customized lingual appliance.

    PubMed

    Sachse, Tina; Schwestka-Polly, Rainer; Flieger, Stefanie; Wiechmann, Dirk

    2012-05-21

    The role of occlusion concerning temporomandibular disorder is still unclear but seems to be the only component of the stomathognathic system dentists are able to change morphologically. The aim of the paper is to describe the orthodontist's approach for transferring and maintaining a therapeutic splint position into permanent occlusion using a fully customized lingual appliance. Fixed acrylic bite planes on lower molars were used to maintain a symptom-free condyle position prior to orthodontic treatment. Silicone impressions of the arches including the fixed bite planes were used for the Incognito laboratory procedure. Two digital setups were made. One setup represents the target occlusion. A second setup including the bite planes was used to fabricate an additional set of lower molar brackets. In the leveling stage all teeth except the lower molars were settled to maintain the therapeutic condyle position. Finally, the fixed bite planes were stepwise removed and molar brackets were replaced to establish the permanent occlusion planned with the first setup. The advantage of an individual lingual appliance consists in the high level of congruence between the fabricated setups and the final clinical result. Both the individual scope for design and the precision of the appliance were vitally important in the treatment of a patient with a functional disorder of the masticatory system.

  5. Pelvic position and movement during hip replacement.

    PubMed

    Grammatopoulos, G; Pandit, H G; da Assunção, R; Taylor, A; McLardy-Smith, P; De Smet, K A; Murray, D W; Gill, H S

    2014-07-01

    The orientation of the acetabular component is influenced not only by the orientation at which the surgeon implants the component, but also the orientation of the pelvis at the time of implantation. Hence, the orientation of the pelvis at set-up and its movement during the operation, are important. During 67 hip replacements, using a validated photogrammetric technique, we measured how three surgeons orientated the patient's pelvis, how much the pelvis moved during surgery, and what effect these had on the final orientation of the acetabular component. Pelvic orientation at set-up, varied widely (mean (± 2, standard deviation (sd))): tilt 8° (2sd ± 32), obliquity -4° (2sd ± 12), rotation -8° (2sd ± 14). Significant differences in pelvic positioning were detected between surgeons (p < 0.001). The mean angular movement of the pelvis between set-up and component implantation was 9° (sd 6). Factors influencing pelvic movement included surgeon, approach (posterior > lateral), procedure (hip resurfacing > total hip replacement) and type of support (p < 0.001). Although, on average, surgeons achieved their desired acetabular component orientation, there was considerable variability (2sd ± 16) in component orientation. We conclude that inconsistency in positioning the patient at set-up and movement of the pelvis during the operation account for much of the variation in acetabular component orientation. Improved methods of positioning and holding the pelvis are required. ©2014 The British Editorial Society of Bone & Joint Surgery.

  6. In vivo dose verification method in catheter based high dose rate brachytherapy.

    PubMed

    Jaselskė, Evelina; Adlienė, Diana; Rudžianskas, Viktoras; Urbonavičius, Benas Gabrielis; Inčiūra, Arturas

    2017-12-01

    In vivo dosimetry is a powerful tool for dose verification in radiotherapy. Its application in high dose rate (HDR) brachytherapy is usually limited to the estimation of gross errors, due to inability of the dosimetry system/ method to record non-uniform dose distribution in steep dose gradient fields close to the radioactive source. In vivo dose verification in interstitial catheter based HDR brachytherapy is crucial since the treatment is performed inserting radioactive source at the certain positions within the catheters that are pre-implanted into the tumour. We propose in vivo dose verification method for this type of brachytherapy treatment which is based on the comparison between experimentally measured and theoretical dose values calculated at well-defined locations corresponding dosemeter positions in the catheter. Dose measurements were performed using TLD 100-H rods (6 mm long, 1 mm diameter) inserted in a certain sequences into additionally pre-implanted dosimetry catheter. The adjustment of dosemeter positioning in the catheter was performed using reconstructed CT scans of patient with pre-implanted catheters. Doses to three Head&Neck and one Breast cancer patient have been measured during several randomly selected treatment fractions. It was found that the average experimental dose error varied from 4.02% to 12.93% during independent in vivo dosimetry control measurements for selected Head&Neck cancer patients and from 7.17% to 8.63% - for Breast cancer patient. Average experimental dose error was below the AAPM recommended margin of 20% and did not exceed the measurement uncertainty of 17.87% estimated for this type of dosemeters. Tendency of slightly increasing average dose error was observed in every following treatment fraction of the same patient. It was linked to the changes of theoretically estimated dosemeter positions due to the possible patient's organ movement between different treatment fractions, since catheter reconstruction was performed for the first treatment fraction only. These findings indicate potential for further average dose error reduction in catheter based brachytherapy by at least 2-3% in the case that catheter locations will be adjusted before each following treatment fraction, however it requires more detailed investigation. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  7. Is identity per se irrelevant? A contrarian view of self-verification effects.

    PubMed

    Gregg, Aiden P

    2009-01-01

    Self-verification theory (SVT) posits that people who hold negative self-views, such as depressive patients, ironically strive to verify that these self-views are correct, by actively seeking out critical feedback or interaction partners who evaluate them unfavorably. Such verification strivings are allegedly directed towards maximizing subjective perceptions of prediction and control. Nonetheless, verification strivings are also alleged to stabilize maladaptive self-perceptions, and thereby hindering therapeutic recovery. Despite the widespread acceptance of SVT, I contend that the evidence for it is weak and circumstantial. In particular, I contend that that most or all major findings cited in support of SVT can be more economically explained in terms of raison oblige theory (ROT). ROT posits that people with negative self-views solicit critical feedback, not because they want it, but because they their self-view inclines them regard it as probative, a necessary condition for considering it worth obtaining. Relevant findings are reviewed and reinterpreted with an emphasis on depression, and some new empirical data reported. (c) 2008 Wiley-Liss, Inc.

  8. Setup deviations for whole-breast radiotherapy with TomoDirect: A comparison of weekly and biweekly image-guided protocols

    NASA Astrophysics Data System (ADS)

    Jung, Jae Hong; Jung, Joo-Young; Bae, Sun Hyun; Moon, Seong Kwon; Cho, Kwang Hwan

    2016-10-01

    The purpose of this study was to compare patient setup deviations for different image-guided protocols (weekly vs. biweekly) that are used in TomoDirect three-dimensional conformal radiotherapy (TD-3DCRT) for whole-breast radiation therapy (WBRT). A total of 138 defined megavoltage computed tomography (MVCT) image sets from 46 breast cancer cases were divided into two groups based on the imaging acquisition times: weekly or biweekly. The mean error, three-dimensional setup displacement error (3D-error), systematic error (Σ), and random error (σ) were calculated for each group. The 3D-errors were 4.29 ± 1.11 mm and 5.02 ± 1.85 mm for the weekly and biweekly groups, respectively; the biweekly error was 14.6% higher than the weekly error. The systematic errors in the roll angle and the x, y, and z directions were 0.48°, 1.72 mm, 2.18 mm, and 1.85 mm for the weekly protocol and 0.21°, 1.24 mm, 1.39 mm, and 1.85 mm for the biweekly protocol. Random errors in the roll angle and the x, y, and z directions were 25.7%, 40.6%, 40.0%, and 40.8% higher in the biweekly group than in the weekly group. For the x, y, and z directions, the distributions of the treatment frequency at less than 5 mm were 98.6%, 91.3%, and 94.2% in the weekly group and 94.2%, 89.9%, and 82.6% in the biweekly group. Moreover, the roll angles with 0 - 1° were 79.7% and 89.9% in the weekly and the biweekly groups, respectively. Overall, the evaluation of setup deviations for the two protocols revealed no significant differences (p > 0.05). Reducing the frequency of MVCT imaging could have promising effects on imaging doses and machine times during treatment. However, the biweekly protocol was associated with increased random setup deviations in the treatment. We have demonstrated a biweekly protocol of TD-3DCRT for WBRT, and we anticipate that our method may provide an alternative approach for considering the uncertainties in the patient setup.

  9. Technical experiences of implementing a wireless tracking and facial biometric verification system for a clinical environment

    NASA Astrophysics Data System (ADS)

    Liu, Brent; Lee, Jasper; Documet, Jorge; Guo, Bing; King, Nelson; Huang, H. K.

    2006-03-01

    By implementing a tracking and verification system, clinical facilities can effectively monitor workflow and heighten information security in today's growing demand towards digital imaging informatics. This paper presents the technical design and implementation experiences encountered during the development of a Location Tracking and Verification System (LTVS) for a clinical environment. LTVS integrates facial biometrics with wireless tracking so that administrators can manage and monitor patient and staff through a web-based application. Implementation challenges fall into three main areas: 1) Development and Integration, 2) Calibration and Optimization of Wi-Fi Tracking System, and 3) Clinical Implementation. An initial prototype LTVS has been implemented within USC's Healthcare Consultation Center II Outpatient Facility, which currently has a fully digital imaging department environment with integrated HIS/RIS/PACS/VR (Voice Recognition).

  10. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Pettersson, N; Murphy, J; Simpson, D

    Purpose: The use of respiratory gating for management of breathing motion during stereotactic body radiation therapy (SBRT) relies on a consistent relationship between the breathing signal and the actual position of the internal target. This relationship was investigated in patients treated for pancreatic cancer. Methods: Four patients with pancreatic cancer undergoing SBRT that had implanted fiducials in the tumor were included in this study. Treatment plans were generated based on the exhale phases (30–70%) from the pre-treatment 4DCT. The margin between the internal target volume (ITV) and the planning target volume was three mm. After patient setup using cone-beam CT,more » simultaneous fluoroscopic imaging and breathing motion monitoring were used during at least three breathing cycles to verify the fiducial position and to optimize the gating window. After treatment, fluoroscopic images were acquired for verification purposes and exported for retrospective analyses. Fiducial positions were determined using a template-matching algorithm. For each dataset, we established a linear relationship between the fiducial position and the anterior-posterior (AP) breathing signal. The relationships before and after treatment were compared and the dose distribution impact evaluated. Results: Seven pre- and post-treatment fluoroscopic pairs were available for fiducial position analyses in the superior-inferior (SI) and left-right (LR) directions, and five in the AP direction. Time between image acquisitions was typically six to eight minutes. An average absolute change of 1.2±0.7 mm (range: 0.1–1.7) of the SI fiducial position relative to the external signal was found. Corresponding numbers for the LR and AP fiducial positions were 0.9±1.0 mm (range: 0.2–3.0) and 0.5±0.4 mm (range: 0.2–1.2), respectively. The dose distribution impact was small in both the ITV and organs-at-risk. Conclusion: The relationship change between fiducial position and external breathing signal has been observed to be about 1 mm in four pancreas SBRT patients, leading to small dose distribution impact. Pettersson and Cervino are funded by a Varian Medical Systems grant.« less

  11. Model-based engineering for medical-device software.

    PubMed

    Ray, Arnab; Jetley, Raoul; Jones, Paul L; Zhang, Yi

    2010-01-01

    This paper demonstrates the benefits of adopting model-based design techniques for engineering medical device software. By using a patient-controlled analgesic (PCA) infusion pump as a candidate medical device, the authors show how using models to capture design information allows for i) fast and efficient construction of executable device prototypes ii) creation of a standard, reusable baseline software architecture for a particular device family, iii) formal verification of the design against safety requirements, and iv) creation of a safety framework that reduces verification costs for future versions of the device software. 1.

  12. Comparing portable computers with bedside computers when administering medications using bedside medication verification.

    PubMed

    Ludwig-Beymer, Patti; Williams, Phillip; Stimac, Ellen

    2012-01-01

    This research examined bedside medication verification administration in 2 adult critical care units, using portable computers and permanent bedside computers. There were no differences in the number of near-miss errors, the time to administer the medications, or nurse perception of ease of medication administration, care of patients, or reliability of technology. The percentage of medications scanned was significantly higher with the use of permanent bedside computers, and nurses using permanent bedside computers were more likely to agree that the computer was always available.

  13. Dosimetry for audit and clinical trials: challenges and requirements

    NASA Astrophysics Data System (ADS)

    Kron, T.; Haworth, A.; Williams, I.

    2013-06-01

    Many important dosimetry audit networks for radiotherapy have their roots in clinical trial quality assurance (QA). In both scenarios it is essential to test two issues: does the treatment plan conform with the clinical requirements and is the plan a reasonable representation of what is actually delivered to a patient throughout their course of treatment. Part of a sound quality program would be an external audit of these issues with verification of the equivalence of plan and treatment typically referred to as a dosimetry audit. The increasing complexity of radiotherapy planning and delivery makes audits challenging. While verification of absolute dose delivered at a reference point was the standard of external dosimetry audits two decades ago this is often deemed inadequate for verification of treatment approaches such as Intensity Modulated Radiation Therapy (IMRT) and Volumetric Modulated Arc Therapy (VMAT). As such, most dosimetry audit networks have successfully introduced more complex tests of dose delivery using anthropomorphic phantoms that can be imaged, planned and treated as a patient would. The new challenge is to adapt this approach to ever more diversified radiotherapy procedures with image guided/adaptive radiotherapy, motion management and brachytherapy being the focus of current research.

  14. A managed clinical network for cardiac services: set-up, operation and impact on patient care.

    PubMed

    Stc Hamilton, Karen E; Sullivan, Frank M; Donnan, Peter T; Taylor, Rex; Ikenwilo, Divine; Scott, Anthony; Baker, Chris; Wyke, Sally

    2005-01-01

    To investigate the set up and operation of a Managed Clinical Network for cardiac services and assess its impact on patient care. This single case study used process evaluation with observational before and after comparison of indicators of quality of care and costs. The study was conducted in Dumfries and Galloway, Scotland and used a three-level framework. Process evaluation of the network set-up and operation through a documentary review of minutes; guidelines and protocols; transcripts of fourteen semi-structured interviews with health service personnel including senior managers, general practitioners, nurses, cardiologists and members of the public. Outcome evaluation of the impact of the network through interrupted time series analysis of clinical data of 202 patients aged less than 76 years admitted to hospital with a confirmed myocardial infarction one-year pre and one-year post, the establishment of the network. The main outcome measures were differences between indicators of quality of care targeted by network protocols. Economic evaluation of the transaction costs of the set-up and operation of the network and the resource costs of the clinical care of the 202 myocardial infarction patients from the time of hospital admission to 6 months post discharge through interrupted time series analysis. The outcome measure was different in National Health Service resource use. Despite early difficulties, the network was successful in bringing together clinicians, patients and managers to redesign services, exhibiting most features of good network management. The role of the energetic lead clinician was crucial, but the network took time to develop and 'bed down'. Its primary "modus operand" was the development of a myocardial infarction pathway and associated protocols. Of sixteen clinical care indicators, two improved significantly following the launch of the network and nine showed improvements, which were not statistically significant. There was no difference in resource use. The Managed Clinical Network made a difference to ways of working, particularly in breaching traditional boundaries and involving the public, and made modest changes in patient care. However, it required a two-year "set-up" period. Managed clinical networks are complex initiatives with an increasing profile in health care policy. This study suggests that they require energetic leadership and improvements are likely to be slow and incremental.

  15. Deformable structure registration of bladder through surface mapping.

    PubMed

    Xiong, Li; Viswanathan, Akila; Stewart, Alexandra J; Haker, Steven; Tempany, Clare M; Chin, Lee M; Cormack, Robert A

    2006-06-01

    Cumulative dose distributions in fractionated radiation therapy depict the dose to normal tissues and therefore may permit an estimation of the risk of normal tissue complications. However, calculation of these distributions is highly challenging because of interfractional changes in the geometry of patient anatomy. This work presents an algorithm for deformable structure registration of the bladder and the verification of the accuracy of the algorithm using phantom and patient data. In this algorithm, the registration process involves conformal mapping of genus zero surfaces using finite element analysis, and guided by three control landmarks. The registration produces a correspondence between fractions of the triangular meshes used to describe the bladder surface. For validation of the algorithm, two types of balloons were inflated gradually to three times their original size, and several computerized tomography (CT) scans were taken during the process. The registration algorithm yielded a local accuracy of 4 mm along the balloon surface. The algorithm was then applied to CT data of patients receiving fractionated high-dose-rate brachytherapy to the vaginal cuff, with the vaginal cylinder in situ. The patients' bladder filling status was intentionally different for each fraction. The three required control landmark points were identified for the bladder based on anatomy. Out of an Institutional Review Board (IRB) approved study of 20 patients, 3 had radiographically identifiable points near the bladder surface that were used for verification of the accuracy of the registration. The verification point as seen in each fraction was compared with its predicted location based on affine as well as deformable registration. Despite the variation in bladder shape and volume, the deformable registration was accurate to 5 mm, consistently outperforming the affine registration. We conclude that the structure registration algorithm presented works with reasonable accuracy and provides a means of calculating cumulative dose distributions.

  16. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Li Xiong; Viswanathan, Akila; Stewart, Alexandra J.

    Cumulative dose distributions in fractionated radiation therapy depict the dose to normal tissues and therefore may permit an estimation of the risk of normal tissue complications. However, calculation of these distributions is highly challenging because of interfractional changes in the geometry of patient anatomy. This work presents an algorithm for deformable structure registration of the bladder and the verification of the accuracy of the algorithm using phantom and patient data. In this algorithm, the registration process involves conformal mapping of genus zero surfaces using finite element analysis, and guided by three control landmarks. The registration produces a correspondence between fractionsmore » of the triangular meshes used to describe the bladder surface. For validation of the algorithm, two types of balloons were inflated gradually to three times their original size, and several computerized tomography (CT) scans were taken during the process. The registration algorithm yielded a local accuracy of 4 mm along the balloon surface. The algorithm was then applied to CT data of patients receiving fractionated high-dose-rate brachytherapy to the vaginal cuff, with the vaginal cylinder in situ. The patients' bladder filling status was intentionally different for each fraction. The three required control landmark points were identified for the bladder based on anatomy. Out of an Institutional Review Board (IRB) approved study of 20 patients, 3 had radiographically identifiable points near the bladder surface that were used for verification of the accuracy of the registration. The verification point as seen in each fraction was compared with its predicted location based on affine as well as deformable registration. Despite the variation in bladder shape and volume, the deformable registration was accurate to 5 mm, consistently outperforming the affine registration. We conclude that the structure registration algorithm presented works with reasonable accuracy and provides a means of calculating cumulative dose distributions.« less

  17. Characteristics of ACS-verified Level I and Level II trauma centers: A study linking trauma center verification review data and the National Trauma Data Bank of the American College of Surgeons Committee on Trauma.

    PubMed

    Shafi, Shahid; Barnes, Sunni; Ahn, Chul; Hemilla, Mark R; Cryer, H Gill; Nathens, Avery; Neal, Melanie; Fildes, John

    2016-10-01

    The Trauma Quality Improvement Project of the American College of Surgeons (ACS) has demonstrated variations in trauma center outcomes despite similar verification status. The purpose of this study was to identify structural characteristics of trauma centers that affect patient outcomes. Trauma registry data on 361,187 patients treated at 222 ACS-verified Level I and Level II trauma centers were obtained from the National Trauma Data Bank of ACS. These data were used to estimate each center's observed-to-expected (O-E) mortality ratio with 95% confidence intervals using multivariate logistic regression analysis. De-identified data on structural characteristics of these trauma centers were obtained from the ACS Verification Review Committee. Centers in the lowest quartile of mortality based on O-E ratio (n = 56) were compared to the rest (n = 166) using Classification and Regression Tree (CART) analysis to identify institutional characteristics independently associated with high-performing centers. Of the 72 structural characteristics explored, only 3 were independently associated with high-performing centers: annual patient visits to the emergency department of fewer than 61,000; proportion of patients on Medicare greater than 20%; and continuing medical education for emergency department physician liaison to the trauma program ranging from 55 and 113 hours annually. Each 5% increase in O-E mortality ratio was associated with an increase in total length of stay of one day (r = 0.25; p < 0.001). Very few structural characteristics of ACS-verified trauma centers are associated with risk-adjusted mortality. Thus, variations in patient outcomes across trauma centers are likely related to variations in clinical practices. Therapeutic study, level III.

  18. Improvement of a uniqueness-and-anonymity-preserving user authentication scheme for connected health care.

    PubMed

    Xie, Qi; Liu, Wenhao; Wang, Shengbao; Han, Lidong; Hu, Bin; Wu, Ting

    2014-09-01

    Patient's privacy-preserving, security and mutual authentication between patient and the medical server are the important mechanism in connected health care applications, such as telecare medical information systems and personally controlled health records systems. In 2013, Wen showed that Das et al.'s scheme is vulnerable to the replay attack, user impersonation attacks and off-line guessing attacks, and then proposed an improved scheme using biometrics, password and smart card to overcome these weaknesses. However, we show that Wen's scheme is still vulnerable to off-line password guessing attacks, does not provide user's anonymity and perfect forward secrecy. Further, we propose an improved scheme to fix these weaknesses, and use the applied pi calculus based formal verification tool ProVerif to prove the security and authentication.

  19. Assessment of three-dimensional setup errors in image-guided pelvic radiotherapy for uterine and cervical cancer using kilovoltage cone-beam computed tomography and its effect on planning target volume margins.

    PubMed

    Patni, Nidhi; Burela, Nagarjuna; Pasricha, Rajesh; Goyal, Jaishree; Soni, Tej Prakash; Kumar, T Senthil; Natarajan, T

    2017-01-01

    To achieve the best possible therapeutic ratio using high-precision techniques (image-guided radiation therapy/volumetric modulated arc therapy [IGRT/VMAT]) of external beam radiation therapy in cases of carcinoma cervix using kilovoltage cone-beam computed tomography (kV-CBCT). One hundred and five patients of gynecological malignancies who were treated with IGRT (IGRT/VMAT) were included in the study. CBCT was done once a week for intensity-modulated radiation therapy and daily in IGRT/VMAT. These images were registered with the planning CT scan images and translational errors were applied and recorded. In all, 2078 CBCT images were studied. The margins of planning target volume were calculated from the variations in the setup. The setup variation was 5.8, 10.3, and 5.6 mm in anteroposterior, superoinferior, and mediolateral direction. This allowed adequate dose delivery to the clinical target volume and the sparing of organ at risks. Daily kV-CBCT is a satisfactory method of accurate patient positioning in treating gynecological cancers with high-precision techniques. This resulted in avoiding geographic miss.

  20. SU-E-T-490: Independent Three-Dimensional (3D) Dose Verification of VMAT/SBRT Using EPID and Cloud Computing

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ding, A; Han, B; Bush, K

    Purpose: Dosimetric verification of VMAT/SBRT is currently performed on one or two planes in a phantom with either film or array detectors. A robust and easy-to-use 3D dosimetric tool has been sought since the advent of conformal radiation therapy. Here we present such a strategy for independent 3D VMAT/SBRT plan verification system by a combined use of EPID and cloud-based Monte Carlo (MC) dose calculation. Methods: The 3D dosimetric verification proceeds in two steps. First, the plan was delivered with a high resolution portable EPID mounted on the gantry, and the EPID-captured gantry-angle-resolved VMAT/SBRT field images were converted into fluencemore » by using the EPID pixel response function derived from MC simulations. The fluence was resampled and used as the input for an in-house developed Amazon cloud-based MC software to reconstruct the 3D dose distribution. The accuracy of the developed 3D dosimetric tool was assessed using a Delta4 phantom with various field sizes (square, circular, rectangular, and irregular MLC fields) and different patient cases. The method was applied to validate VMAT/SBRT plans using WFF and FFF photon beams (Varian TrueBeam STX). Results: It was found that the proposed method yielded results consistent with the Delta4 measurements. For points on the two detector planes, a good agreement within 1.5% were found for all the testing fields. Patient VMAT/SBRT plan studies revealed similar level of accuracy: an average γ-index passing rate of 99.2± 0.6% (3mm/3%), 97.4± 2.4% (2mm/2%), and 72.6± 8.4 % ( 1mm/1%). Conclusion: A valuable 3D dosimetric verification strategy has been developed for VMAT/SBRT plan validation. The technique provides a viable solution for a number of intractable dosimetry problems, such as small fields and plans with high dose gradient.« less

  1. Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218.

    PubMed

    Miften, Moyed; Olch, Arthur; Mihailidis, Dimitris; Moran, Jean; Pawlicki, Todd; Molineu, Andrea; Li, Harold; Wijesooriya, Krishni; Shi, Jie; Xia, Ping; Papanikolaou, Nikos; Low, Daniel A

    2018-04-01

    Patient-specific IMRT QA measurements are important components of processes designed to identify discrepancies between calculated and delivered radiation doses. Discrepancy tolerance limits are neither well defined nor consistently applied across centers. The AAPM TG-218 report provides a comprehensive review aimed at improving the understanding and consistency of these processes as well as recommendations for methodologies and tolerance limits in patient-specific IMRT QA. The performance of the dose difference/distance-to-agreement (DTA) and γ dose distribution comparison metrics are investigated. Measurement methods are reviewed and followed by a discussion of the pros and cons of each. Methodologies for absolute dose verification are discussed and new IMRT QA verification tools are presented. Literature on the expected or achievable agreement between measurements and calculations for different types of planning and delivery systems are reviewed and analyzed. Tests of vendor implementations of the γ verification algorithm employing benchmark cases are presented. Operational shortcomings that can reduce the γ tool accuracy and subsequent effectiveness for IMRT QA are described. Practical considerations including spatial resolution, normalization, dose threshold, and data interpretation are discussed. Published data on IMRT QA and the clinical experience of the group members are used to develop guidelines and recommendations on tolerance and action limits for IMRT QA. Steps to check failed IMRT QA plans are outlined. Recommendations on delivery methods, data interpretation, dose normalization, the use of γ analysis routines and choice of tolerance limits for IMRT QA are made with focus on detecting differences between calculated and measured doses via the use of robust analysis methods and an in-depth understanding of IMRT verification metrics. The recommendations are intended to improve the IMRT QA process and establish consistent, and comparable IMRT QA criteria among institutions. © 2018 American Association of Physicists in Medicine.

  2. Real-time in vivo dosimetry with MOSFET detectors in serial tomotherapy for head and neck cancer patients.

    PubMed

    Qi, Zhen-Yu; Deng, Xiao-Wu; Huang, Shao-Min; Shiu, Almon; Lerch, Michael; Metcalfe, Peter; Rosenfeld, Anatoly; Kron, Tomas

    2011-08-01

    A real-time dose verification method using a recently designed metal oxide semiconductor field effect transistor (MOSFET) dosimetry system was evaluated for quality assurance (QA) of intensity-modulated radiation therapy (IMRT). Following the investigation of key parameters that might affect the accuracy of MOSFET measurements (i.e., source surface distance [SSD], field size, beam incident angles and radiation energy spectrum), the feasibility of this detector in IMRT dose verification was demonstrated by comparison with ion chamber measurements taken in an IMRT QA phantom. Real-time in vivo measurements were also performed with the MOSFET system during serial tomotherapy treatments administered to 8 head and neck cancer patients. MOSFET sensitivity did not change with SSD. For field sizes smaller than 20 × 20 cm(2), MOFET sensitivity varied within 1.0%. The detector angular response was isotropic within 2% over 360°, and the observed sensitivity variation due to changes in the energy spectrum was negligible in 6-MV photons. MOSFET system measurements and ion chamber measurements agreed at all points in IMRT phantom plan verification, within 5%. The mean difference between 48 IMRT MOSFET-measured doses and calculated values in 8 patients was 3.33% and ranged from -2.20% to 7.89%. More than 90% of the total measurements had deviations of less than 5% from the planned doses. The MOSFET dosimetry system has been proven to be an effective tool in evaluating the actual dose within individual patients during IMRT treatment. Copyright © 2011 Elsevier Inc. All rights reserved.

  3. Target coverage in image-guided stereotactic body radiotherapy of liver tumors.

    PubMed

    Wunderink, Wouter; Méndez Romero, Alejandra; Vásquez Osorio, Eliana M; de Boer, Hans C J; Brandwijk, René P; Levendag, Peter C; Heijmen, Ben J M

    2007-05-01

    To determine the effect of image-guided procedures (with computed tomography [CT] and electronic portal images before each treatment fraction) on target coverage in stereotactic body radiotherapy for liver patients using a stereotactic body frame (SBF) and abdominal compression. CT guidance was used to correct for day-to-day variations in the tumor's mean position in the SBF. By retrospectively evaluating 57 treatment sessions, tumor coverage, as obtained with the clinically applied CT-guided protocol, was compared with that of alternative procedures. The internal target volume-plus (ITV(+)) was introduced to explicitly include uncertainties in tumor delineations resulting from CT-imaging artifacts caused by residual respiratory motion. Tumor coverage was defined as the volume overlap of the ITV(+), derived from a tumor delineated in a treatment CT scan, and the planning target volume. Patient stability in the SBF, after acquisition of the treatment CT scan, was evaluated by measuring the displacement of the bony anatomy in the electronic portal images relative to CT. Application of our clinical protocol (with setup corrections following from manual measurements of the distances between the contours of the planning target volume and the daily clinical target volume in three orthogonal planes, multiple two-dimensional) increased the frequency of nearly full (> or = 99%) ITV(+) coverage to 77% compared with 63% without setup correction. An automated three-dimensional method further improved the frequency to 96%. Patient displacements in the SBF were generally small (< or = 2 mm, 1 standard deviation), but large craniocaudal displacements (maximal 7.2 mm) were occasionally observed. Daily, CT-assisted patient setup may substantially improve tumor coverage, especially with the automated three-dimensional procedure. In the present treatment design, patient stability in the SBF should be verified with portal imaging.

  4. Accuracy and Consistency of Respiratory Gating in Abdominal Cancer Patients

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ge, Jiajia; Santanam, Lakshmi; Yang, Deshan

    2013-03-01

    Purpose: To evaluate respiratory gating accuracy and intrafractional consistency for abdominal cancer patients treated with respiratory gated treatment on a regular linear accelerator system. Methods and Materials: Twelve abdominal patients implanted with fiducials were treated with amplitude-based respiratory-gated radiation therapy. On the basis of daily orthogonal fluoroscopy, the operator readjusted the couch position and gating window such that the fiducial was within a setup margin (fiducial-planning target volume [f-PTV]) when RPM indicated “beam-ON.” Fifty-five pre- and post-treatment fluoroscopic movie pairs with synchronized respiratory gating signal were recorded. Fiducial motion traces were extracted from the fluoroscopic movies using a template matchingmore » algorithm and correlated with f-PTV by registering the digitally reconstructed radiographs with the fluoroscopic movies. Treatment was determined to be “accurate” if 50% of the fiducial area stayed within f-PTV while beam-ON. For movie pairs that lost gating accuracy, a MATLAB program was used to assess whether the gating window was optimized, the external-internal correlation (EIC) changed, or the patient moved between movies. A series of safety margins from 0.5 mm to 3 mm was added to f-PTV for reassessing gating accuracy. Results: A decrease in gating accuracy was observed in 44% of movie pairs from daily fluoroscopic movies of 12 abdominal patients. Three main causes for inaccurate gating were identified as change of global EIC over time (∼43%), suboptimal gating setup (∼37%), and imperfect EIC within movie (∼13%). Conclusions: Inconsistent respiratory gating accuracy may occur within 1 treatment session even with a daily adjusted gating window. To improve or maintain gating accuracy during treatment, we suggest using at least a 2.5-mm safety margin to account for gating and setup uncertainties.« less

  5. One lens optical correlation: application to face recognition.

    PubMed

    Jridi, Maher; Napoléon, Thibault; Alfalou, Ayman

    2018-03-20

    Despite its extensive use, the traditional 4f Vander Lugt Correlator optical setup can be further simplified. We propose a lightweight correlation scheme where the decision is taken in the Fourier plane. For this purpose, the Fourier plane is adapted and used as a decision plane. Then, the offline phase and the decision metric are re-examined in order to keep a reasonable recognition rate. The benefits of the proposed approach are numerous: (1) it overcomes the constraints related to the use of a second lens; (2) the optical correlation setup is simplified; (3) the multiplication with the correlation filter can be done digitally, which offers a higher adaptability according to the application. Moreover, the digital counterpart of the correlation scheme is lightened since with the proposed scheme we get rid of the inverse Fourier transform (IFT) calculation (i.e., decision directly in the Fourier domain without resorting to IFT). To assess the performance of the proposed approach, an insight into digital hardware resources saving is provided. The proposed method involves nearly 100 times fewer arithmetic operators. Moreover, from experimental results in the context of face verification-based correlation, we demonstrate that the proposed scheme provides comparable or better accuracy than the traditional method. One interesting feature of the proposed scheme is that it could greatly outperform the traditional scheme for face identification application in terms of sensitivity to face orientation. The proposed method is found to be digital/optical implementation-friendly, which facilitates its integration on a very broad range of scenarios.

  6. Single Vocal Cord Irradiation: Image Guided Intensity Modulated Hypofractionated Radiation Therapy for T1a Glottic Cancer: Early Clinical Results

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Al-Mamgani, Abrahim, E-mail: a.almamgani@nki.nl; Kwa, Stefan L.S.; Tans, Lisa

    2015-10-01

    Purpose: To report, from a retrospective analysis of prospectively collected data, on the feasibility, outcome, toxicity, and voice-handicap index (VHI) of patients with T1a glottic cancer treated by a novel intensity modulated radiation therapy technique developed at our institution to treat only the involved vocal cord: single vocal cord irradiation (SVCI). Methods and Materials: Thirty patients with T1a glottic cancer were treated by means of SVCI. Dose prescription was set to 16 × 3.63 Gy (total dose 58.08 Gy). The clinical target volume was the entire vocal cord. Setup verification was done by means of an online correction protocol using cone beam computed tomography. Datamore » for voice quality assessment were collected prospectively at baseline, end of treatment, and 4, 6, and 12 weeks and 6, 12, and 18 months after treatment using VHI questionnaires. Results: After a median follow-up of 30 months (range, 7-50 months), the 2-year local control and overall survival rates were 100% and 90% because no single local recurrence was reported and 3 patients died because of comorbidity. All patients have completed the intended treatment schedule; no treatment interruptions and no grade 3 acute toxicity were reported. Grade 2 acute dermatitis or dysphagia was reported in only 5 patients (17%). No serious late toxicity was reported; only 1 patient developed temporary grade 2 laryngeal edema, and responded to a short-course of corticosteroid. The VHI improved significantly, from 33.5 at baseline to 9.5 and 10 at 6 weeks and 18 months, respectively (P<.001). The control group, treated to the whole larynx, had comparable local control rates (92.2% vs 100%, P=.24) but more acute toxicity (66% vs 17%, P<.0001) and higher VHI scores (23.8 and 16.7 at 6 weeks and 18 months, respectively, P<.0001). Conclusion: Single vocal cord irradiation is feasible and resulted in maximal local control rate at 2 years. The deterioration in VHI scores was slight and temporary and subsequently improved to normal levels. Long-term follow-up is needed to consolidate these promising results.« less

  7. SU-F-BRE-13: Replacing Pre-Treatment Phantom QA with 3D In-Vivo Portal Dosimetry for IMRT Breast Cancer

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Stroom, J; Vieira, S; Greco, C

    Purpose: Pre-treatment QA of individual treatment plans requires costly linac time and physics effort. Starting with IMRT breast treatments, we aim to replace pre-treatment QA with in-vivo portal dosimetry. Methods: Our IMRT breast cancer plans are routinely measured using the ArcCheck device (SunNuclear). 2D-Gamma analysis is performed with 3%/3mm criteria and the percentage of points with gamma<1 (nG1) is calculated within the 50% isodose surface. Following AAPM recommendations, plans with nG1<90% are approved; others need further inspection and might be rejected. For this study, we used invivo portal dosimetry (IPD) to measure the 3D back-projected dose of the first threemore » fractions for IMRT breast plans. Patient setup was online corrected before for all measured fractions. To reduce patient related uncertainties, the three IPD results were averaged and 3D-gamma analysis was applied with abovementioned criteria . For a subset of patients, phantom portal dosimetry (PPD) was also performed on a slab phantom. Results: Forty consecutive breast patients with plans that fitted the EPID were analysed. The average difference between planned and IPD dose in the reference point was −0.7+/−1.6% (1SD). Variation in nG1 between the 3 invivo fractions was about 6% (1SD). The average nG1 for IPD was 89+/−6%, worse than ArcCheck (95+/−3%). This can be explained by patient related factors such as changes in anatomy and/or model deficiencies due to e.g. inhomogeneities. For the 20 cases with PPD, mean nG1 was equal to ArcCheck values, which indicates that the two systems are equally accurate. These data therefore suggest that proper criteria for 3D invivo verification of breast treatments should be nG1>80% instead of nG1>90%, which, for our breast cases, would result in 5% (2/40) further inspections. Conclusion: First-fraction in-vivo portal dosimetry using new gamma-evaluation criteria will replace phantom measurements in our institution, saving resources and yielding 3D dosimetry of the actual patient treatment.« less

  8. Modelling and experimental verification of a water alleviation system for the NASP. [National Aerospace Plane

    NASA Technical Reports Server (NTRS)

    Vanfossen, G. James

    1992-01-01

    One possible low speed propulsion system for the National Aerospace Plane is a liquid air cycle engine (LACE). The LACE system uses the heat sink in the liquid hydrogen propellant to liquefy air in a heat exchanger which is then pumped up to high pressure and used as the oxidizer in a hydrogen liquid air rocket. The inlet airstream must be dehumidified or moisture could freeze on the cryogenic heat exchangers and block them. The main objective of this research has been to develop a computer simulation of the cold tube/antifreeze-spray water alleviation system and to verify the model with experimental data. An experimental facility has been built and humid air tests were conducted on a generic heat exchanger to obtain condensing data for code development. The paper describes the experimental setup, outlines the method of calculation used in the code, and presents comparisons of the calculations and measurements. Cause of discrepancies between the model and data are explained.

  9. Novel Hyperspectral Sun Photometer for Satellite Remote Sensing Data Radiometeic Calibration and Atmospheric Aerosol Studies

    NASA Technical Reports Server (NTRS)

    Pagnutti, Mary; Ryan, Robert E.; Holekamp, Kara; Harrington, Gary; Frisbie, Troy

    2006-01-01

    A simple and cost-effective, hyperspectral sun photometer for radiometric vicarious remote sensing system calibration, air quality monitoring, and potentially in-situ planetary climatological studies, was developed. The device was constructed solely from off the shelf components and was designed to be easily deployable for support of short-term verification and validation data collects. This sun photometer not only provides the same data products as existing multi-band sun photometers but also the potential of hyperspectral optical depth and diffuse-to-global products. As compared to traditional sun photometers, this device requires a simpler setup, less data acquisition time and allows for a more direct calibration approach. Fielding this instrument has also enabled Stennis Space Center (SSC) Applied Sciences Directorate personnel to cross-calibrate existing sun photometers. This innovative research will position SSC personnel to perform air quality assessments in support of the NASA Applied Sciences Program's National Applications program element as well as to develop techniques to evaluate aerosols in a Martian or other planetary atmosphere.

  10. Optical information authentication using compressed double-random-phase-encoded images and quick-response codes.

    PubMed

    Wang, Xiaogang; Chen, Wen; Chen, Xudong

    2015-03-09

    In this paper, we develop a new optical information authentication system based on compressed double-random-phase-encoded images and quick-response (QR) codes, where the parameters of optical lightwave are used as keys for optical decryption and the QR code is a key for verification. An input image attached with QR code is first optically encoded in a simplified double random phase encoding (DRPE) scheme without using interferometric setup. From the single encoded intensity pattern recorded by a CCD camera, a compressed double-random-phase-encoded image, i.e., the sparse phase distribution used for optical decryption, is generated by using an iterative phase retrieval technique with QR code. We compare this technique to the other two methods proposed in literature, i.e., Fresnel domain information authentication based on the classical DRPE with holographic technique and information authentication based on DRPE and phase retrieval algorithm. Simulation results show that QR codes are effective on improving the security and data sparsity of optical information encryption and authentication system.

  11. Overcoming correlation fluctuations in two-photon interference experiments with differently bright and independently blinking remote quantum emitters

    NASA Astrophysics Data System (ADS)

    Weber, Jonas H.; Kettler, Jan; Vural, Hüseyin; Müller, Markus; Maisch, Julian; Jetter, Michael; Portalupi, Simone L.; Michler, Peter

    2018-05-01

    As a fundamental building block for quantum computation and communication protocols, the correct verification of the two-photon interference (TPI) contrast between two independent quantum light sources is of utmost importance. Here, we experimentally demonstrate how frequently present blinking dynamics and changes in emitter brightness critically affect the Hong-Ou-Mandel-type (HOM) correlation histograms of remote TPI experiments measured via the commonly utilized setup configuration. We further exploit this qualitative and quantitative explanation of the observed correlation dynamics to establish an alternative interferometer configuration, which is overcoming the discussed temporal fluctuations, giving rise to an error-free determination of the remote TPI visibility. We prove full knowledge of the obtained correlation by reproducing the measured correlation statistics via Monte Carlo simulations. As an exemplary system, we make use of two pairs of remote semiconductor quantum dots; however, the same conclusions apply for TPI experiments with flying qubits from any kind of remote solid-state quantum emitters.

  12. Design, fabrication, and verification of a three-dimensional autocollimator.

    PubMed

    Yin, Yanhe; Cai, Sheng; Qiao, Yanfeng

    2016-12-10

    The autocollimator is an optical instrument for noncontact angle measurement with high resolution and a long detection range. It measures two-dimensional angles, i.e., pitch and yaw, but not roll. In this paper, we present a novelly structured autocollimator capable of measuring three-dimensional (3D) angles simultaneously. In this setup, two collimated beams of different wavelengths are projected onto a right-angle prism. One beam is reflected by the hypotenuse of the prism and received by an autocollimation unit for detecting pitch and yaw. The other is reflected by the two legs of the right-angle prism and received by a moiré fringe imaging unit for detecting roll. Furthermore, a prototype is designed and fabricated. Experiments are carried out to evaluate its basic performance. Calibration results show that this prototype has angular RMS errors of less than 5 arcsec in all 3Ds over a range of 1000 arcsec at a working distance of 2 m.

  13. Thickness determination of polymeric multilayer surface protection systems for concrete by means of pulse thermography

    NASA Astrophysics Data System (ADS)

    Altenburg, S. J.; Krankenhagen, R.; Bavendiek, F.

    2017-02-01

    For thickness determination of polymer based surface protection systems for concrete surfaces, so far only destructive measurement techniques are available. Pulse thermography appears to be well suited for non-destructive thickness evaluation in these systems. Here, we present first results of the development of a respective measurement and analysis procedure. Since surface protection systems consist of a number of layers, a model for the calculation of the surface temperature of a multi-layer structure on a semi-infinite (concrete) substrate in pulse thermography setup was developed. It considers semitransparency of the upmost layer and thermal losses at the surface. It also supports the use of an arbitrary temporal shape of the heating pulse to properly describe the measurement conditions for different heat sources. Simulations for one and three layers on the substrate are presented and first results from fitting the model to experimental data for thickness determination and verification of the model are presented.

  14. Random technique to encode complex valued holograms with on axis reconstruction onto phase-only displays.

    PubMed

    Luis Martínez Fuentes, Jose; Moreno, Ignacio

    2018-03-05

    A new technique for encoding the amplitude and phase of diffracted fields in digital holography is proposed. It is based on a random spatial multiplexing of two phase-only diffractive patterns. The first one is the phase information of the intended pattern, while the second one is a diverging optical element whose purpose is the control of the amplitude. A random number determines the choice between these two diffractive patterns at each pixel, and the amplitude information of the desired field governs its discrimination threshold. This proposed technique is computationally fast and does not require iterative methods, and the complex field reconstruction appears on axis. We experimentally demonstrate this new encoding technique with holograms implemented onto a flicker-free phase-only spatial light modulator (SLM), which allows the axial generation of such holograms. The experimental verification includes the phase measurement of generated patterns with a phase-shifting polarization interferometer implemented in the same experimental setup.

  15. Adaptive correction to the speckle correlation fringes by using a twisted-nematic liquid-crystal display.

    PubMed

    Hack, Erwin; Gundu, Phanindra Narayan; Rastogi, Pramod

    2005-05-10

    An innovative technique for reducing speckle noise and improving the intensity profile of the speckle correlation fringes is presented. The method is based on reducing the range of the modulation intensity values of the speckle interference pattern. After the fringe pattern is corrected adaptively at each pixel, a simple morphological filtering of the fringes is sufficient to obtain smoothed fringes. The concept is presented both analytically and by simulation by using computer-generated speckle patterns. The experimental verification is performed by using an amplitude-only spatial light modulator (SLM) in a conventional electronic speckle pattern interferometry setup. The optical arrangement for tuning a commercially available LCD array for amplitude-only behavior is described. The method of feedback to the LCD SLM to modulate the intensity of the reference beam in order to reduce the modulation intensity values is explained, and the resulting fringe pattern and increase in the signal-to-noise ratio are discussed.

  16. Experimental verification of an indefinite causal order

    PubMed Central

    Rubino, Giulia; Rozema, Lee A.; Feix, Adrien; Araújo, Mateus; Zeuner, Jonas M.; Procopio, Lorenzo M.; Brukner, Časlav; Walther, Philip

    2017-01-01

    Investigating the role of causal order in quantum mechanics has recently revealed that the causal relations of events may not be a priori well defined in quantum theory. Although this has triggered a growing interest on the theoretical side, creating processes without a causal order is an experimental task. We report the first decisive demonstration of a process with an indefinite causal order. To do this, we quantify how incompatible our setup is with a definite causal order by measuring a “causal witness.” This mathematical object incorporates a series of measurements that are designed to yield a certain outcome only if the process under examination is not consistent with any well-defined causal order. In our experiment, we perform a measurement in a superposition of causal orders—without destroying the coherence—to acquire information both inside and outside of a “causally nonordered process.” Using this information, we experimentally determine a causal witness, demonstrating by almost 7 SDs that the experimentally implemented process does not have a definite causal order. PMID:28378018

  17. Limitations in learning: How treatment verifications fail and what to do about it?

    PubMed

    Richardson, Susan; Thomadsen, Bruce

    The purposes of this study were: to provide dialog on why classic incident learning systems have been insufficient for patient safety improvements, discuss failures in treatment verification, and to provide context to the reasons and lessons that can be learned from these failures. Historically, incident learning in brachytherapy is performed via database mining which might include reading of event reports and incidents followed by incorporating verification procedures to prevent similar incidents. A description of both classic event reporting databases and current incident learning and reporting systems is given. Real examples of treatment failures based on firsthand knowledge are presented to evaluate the effectiveness of verification. These failures will be described and analyzed by outlining potential pitfalls and problems based on firsthand knowledge. Databases and incident learning systems can be limited in value and fail to provide enough detail for physicists seeking process improvement. Four examples of treatment verification failures experienced firsthand by experienced brachytherapy physicists are described. These include both underverification and oververification of various treatment processes. Database mining is an insufficient method to affect substantial improvements in the practice of brachytherapy. New incident learning systems are still immature and being tested. Instead, a new method of shared learning and implementation of changes must be created. Copyright © 2017 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  18. Accuracy of the dose-shift approximation in estimating the delivered dose in SBRT of lung tumors considering setup errors and breathing motions.

    PubMed

    Karlsson, Kristin; Lax, Ingmar; Lindbäck, Elias; Poludniowski, Gavin

    2017-09-01

    Geometrical uncertainties can result in a delivered dose to the tumor different from that estimated in the static treatment plan. The purpose of this project was to investigate the accuracy of the dose calculated to the clinical target volume (CTV) with the dose-shift approximation, in stereotactic body radiation therapy (SBRT) of lung tumors considering setup errors and breathing motion. The dose-shift method was compared with a beam-shift method with dose recalculation. Included were 10 patients (10 tumors) selected to represent a variety of SBRT-treated lung tumors in terms of tumor location, CTV volume, and tumor density. An in-house developed toolkit within a treatment planning system allowed the shift of either the dose matrix or a shift of the beam isocenter with dose recalculation, to simulate setup errors and breathing motion. Setup shifts of different magnitudes (up to 10 mm) and directions as well as breathing with different peak-to-peak amplitudes (up to 10:5:5 mm) were modeled. The resulting dose-volume histograms (DVHs) were recorded and dose statistics were extracted. Generally, both the dose-shift and beam-shift methods resulted in calculated doses lower than the static planned dose, although the minimum (D 98% ) dose exceeded the prescribed dose in all cases, for setup shifts up to 5 mm. The dose-shift method also generally underestimated the dose compared with the beam-shift method. For clinically realistic systematic displacements of less than 5 mm, the results demonstrated that in the minimum dose region within the CTV, the dose-shift method was accurate to 2% (root-mean-square error). Breathing motion only marginally degraded the dose distributions. Averaged over the patients and shift directions, the dose-shift approximation was determined to be accurate to approximately 2% (RMS) within the CTV, for clinically relevant geometrical uncertainties for SBRT of lung tumors.

  19. A long arm for ultrasound: a combined robotic focused ultrasound setup for magnetic resonance-guided focused ultrasound surgery.

    PubMed

    Krafft, Axel J; Jenne, Jürgen W; Maier, Florian; Stafford, R Jason; Huber, Peter E; Semmler, Wolfhard; Bock, Michael

    2010-05-01

    Focused ultrasound surgery (FUS) is a highly precise noninvasive procedure to ablate pathogenic tissue. FUS therapy is often combined with magnetic resonance (MR) imaging as MR imaging offers excellent target identification and allows for continuous monitoring of FUS induced temperature changes. As the dimensions of the ultrasound (US) focus are typically much smaller than the targeted volume, multiple sonications and focus repositioning are interleaved to scan the focus over the target volume. Focal scanning can be achieved electronically by using phased-array US transducers or mechanically by using dedicated mechanical actuators. In this study, the authors propose and evaluate the precision of a combined robotic FUS setup to overcome some of the limitations of the existing MRgFUS systems. Such systems are typically integrated into the patient table of the MR scanner and thus only provide an application of the US wave within a limited spatial range from below the patient. The fully MR-compatible robotic assistance system InnoMotion (InnoMedic GmbH, Herxheim, Germany) was originally designed for MR-guided interventions with needles. It offers five pneumatically driven degrees of freedom and can be moved over a wide range within the bore of the magnet. In this work, the robotic system was combined with a fixed-focus US transducer (frequency: 1.7 MHz; focal length: 68 mm, and numerical aperture: 0.44) that was integrated into a dedicated, in-house developed treatment unit for FUS application. A series of MR-guided focal scanning procedures was performed in a polyacrylamide-egg white gel phantom to assess the positioning accuracy of the combined FUS setup. In animal experiments with a 3-month-old domestic pig, the system's potential and suitability for MRgFUS was tested. In phantom experiments, a total targeting precision of about 3 mm was found, which is comparable to that of the existing MRgFUS systems. Focus positioning could be performed within a few seconds. During in vivo experiments, a defined pattern of single thermal lesions and a therapeutically relevant confluent thermal lesion could be created. The creation of local tissue necrosis by coagulation was confirmed by post-FUS MR imaging and histological examinations on the treated tissue sample. During all sonications in phantom and in vivo, reliable MR imaging and online MR thermometry could be performed without compromises due to operation of the combined robotic FUS setup. Compared to the existing MRgFUS systems, the combined robotic FUS approach offers a wide range of spatial flexibility so that highly flexible application of the US wave would be possible, for example, to avoid risk structures within the US field. The setup might help to realize new ways of patient access in MRgFUS therapy. The setup is compatible with any closed-bore MR system and does not require an especially designed patient table.

  20. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Toltz, A; Seuntjens, J; Hoesl, M

    Purpose: With the aim of reducing acute esophageal radiation toxicity in pediatric patients receiving craniospinal irradiation (CSI), we investigated the implementation of an in-vivo, adaptive proton therapy range verification methodology. Simulation experiments and in-phantom measurements were conducted to validate the range verification technique for this clinical application. Methods: A silicon diode array system has been developed and experimentally tested in phantom for passively scattered proton beam range verification for a prostate treatment case by correlating properties of the detector signal to the water equivalent path length (WEPL). We propose to extend the methodology to verify range distal to the vertebralmore » body for pediatric CSI cases by placing this small volume dosimeter in the esophagus of the anesthetized patient immediately prior to treatment. A set of calibration measurements was performed to establish a time signal to WEPL fit for a “scout” beam in a solid water phantom. Measurements are compared against Monte Carlo simulation in GEANT4 using the Tool for Particle Simulation (TOPAS). Results: Measurements with the diode array in a spread out Bragg peak of 14 cm modulation width and 15 cm range (177 MeV passively scattered beam) in solid water were successfully validated against proton fluence rate simulations in TOPAS. The resulting calibration curve allows for a sensitivity analysis of detector system response with dose rate in simulation and with individual diode position through simulation on patient CT data. Conclusion: Feasibility has been shown for the application of this range verification methodology to pediatric CSI. An in-vivo measurement to determine the WEPL to the inner surface of the esophagus will allow for personalized adjustment of the treatment plan to ensure sparing of the esophagus while confirming target coverage. A Toltz acknowledges partial support by the CREATE Medical Physics Research Training Network grant of the Natural Sciences and Engineering Research Council (Grant number: 432290)« less

  1. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Lucconi, G; Department of Radiation Oncology, Massachusetts General Hospital, Boston, MA; Bentefour, E

    Purpose: The clinical commissioning of a workflow for pre-treatment range verification/adjustment for the head treatment of pediatric medulloblastoma patients, including dose monitoring during treatment. Methods: An array of Si-diodes (DIODES Incorporated) is placed on the patient skin on the opposite side to the beam entrance. A “scout” SOBP beam, with a longer beam range to cover the diodes in its plateau, is delivered; the measured signal is analyzed and the extracted water equivalent path lengths (WEPL) are compared to the expected values, revealing if a range correction is needed. Diodes stay in place during treatment to measure dose. The workflowmore » was tested in solid water and head phantoms and validated against independent WEPL measurements. Both measured WEPL and skin doses were compared to computed values from the TPS (XiO); a Markus chamber was used for reference dose measurements. Results: The WEPL accuracy of the method was verified by comparing it with the dose extinction method. It resulted, for both solid water and head phantom, in the sub-millimeter range, with a deviation less than 1% to the value extracted from the TPS. The accuracy of dose measurements in the fall-off part of the dose profile was validated against the Markus chamber. The entire range verification workflow was successfully tested for the mock-treatment of head phantom with the standard delivery of 90 cGy per field per fraction. The WEPL measurement revealed no need for range correction. The dose measurements agreed to better than 4% with the prescription dose. The robustness of the method and workflow, including detector array, hardware set and software functions, was successfully stress-tested with multiple repetitions. Conclusion: The performance of the in-vivo range verification system and related workflow meet the clinical requirements in terms of the needed WEPL accuracy for pretreatment range verification with acceptable dose to the patient.« less

  2. An ontology based trust verification of software license agreement

    NASA Astrophysics Data System (ADS)

    Lu, Wenhuan; Li, Xiaoqing; Gan, Zengqin; Wei, Jianguo

    2017-08-01

    When we install software or download software, there will show up so big mass document to state the rights and obligations, for which lots of person are not patient to read it or understand it. That would may make users feel distrust for the software. In this paper, we propose an ontology based verification for Software License Agreement. First of all, this work proposed an ontology model for domain of Software License Agreement. The domain ontology is constructed by proposed methodology according to copyright laws and 30 software license agreements. The License Ontology can act as a part of generalized copyright law knowledge model, and also can work as visualization of software licenses. Based on this proposed ontology, a software license oriented text summarization approach is proposed which performances showing that it can improve the accuracy of software licenses summarizing. Based on the summarization, the underline purpose of the software license can be explicitly explored for trust verification.

  3. SU-F-T-680: Radiobiological Analysis of the Impact of Daily Patient Deformation and Setup Variations Through the Use of the Cone Beam CT and Deformable Image Registration in Lung Cancer IMRT

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jurkovic, I; Stathakis, S; Markovic, M

    Purpose: To estimate the dose distributions delivered to the patient in each treatment fraction using deformable image registration (DIR) and assess the radiobiological impact of the inter-fraction variations due to patient deformation and setup. Methods: The work is based on the cone beam CT (CBCT) images and treatment plans of two lung cancer patients. Both patients were treated with intensity modulated radiation therapy (IMRT) to 66Gy in 2Gy/fraction. The treatment plans were exported from the treatment planning system (TPS) to the Velocity AI where DIR was performed and the same deformation matrix was used for the deformation of the plannedmore » dose distribution and organ contours to each CBCT dataset. A radiobiological analysis was performed based on the radiobiological parameters of the involved organs at risk (OARs) and planning target volume (PTV). Using the complication free tumor control probability (P+) index, differences in P+ were observed between each CBCT as well as between CBCT and planning dose distributions. Results: The optimal CBCT P? values ranged from 91.6 % to 94.8 % for patient #1 and from 88.8 % to 90.6 % for patient #2. At the dose level of the clinical prescription, the CBCT P+ values ranged from 80.3% to 80.7% for patient #1 and from 80.7% to 81.0% for the patient #2. The planning CT P+ values were 81.0% and 80.7% for the two patients, respectively. These differences emphasize the significance of using the radiobiological analysis when assessing changes in the dose distribution due to the tumor motion and lung deformations. Conclusion: Daily setup variations yield to differences in the actual dose delivered versus the planned one. The observed differences were rather small when only looking at the dosimetric comparison of the dose distributions, however the radiobiology analysis was able to detect clinically relevant differences among the studied dose distributions.« less

  4. Facts and Stats - Hydrocephalus Association

    MedlinePlus

    ... this in Theme Options SIGN IN File Download Verification Thank you for downloading our publication! We will ... I am ...: Child Friend Grandchild Grandparent Medical Professional Self Sibling Parent Spouse Other Patient Diagnosed Age: Infants ( ...

  5. Test of Compton camera components for prompt gamma imaging at the ELBE bremsstrahlung beam

    NASA Astrophysics Data System (ADS)

    Hueso-González, F.; Golnik, C.; Berthel, M.; Dreyer, A.; Enghardt, W.; Fiedler, F.; Heidel, K.; Kormoll, T.; Rohling, H.; Schöne, S.; Schwengner, R.; Wagner, A.; Pausch, G.

    2014-05-01

    In the context of ion beam therapy, particle range verification is a major challenge for the quality assurance of the treatment. One approach is the measurement of the prompt gamma rays resulting from the tissue irradiation. A Compton camera based on several position sensitive gamma ray detectors, together with an imaging algorithm, is expected to reconstruct the prompt gamma ray emission density map, which is correlated with the dose distribution. At OncoRay and Helmholtz-Zentrum Dresden-Rossendorf (HZDR), a Compton camera setup is being developed consisting of two scatter planes: two CdZnTe (CZT) cross strip detectors, and an absorber consisting of one Lu2SiO5 (LSO) block detector. The data acquisition is based on VME electronics and handled by software developed on the ROOT framework. The setup has been tested at the linear electron accelerator ELBE at HZDR, which is used in this experiment to produce bunched bremsstrahlung photons with up to 12.5 MeV energy and a repetition rate of 13 MHz. Their spectrum has similarities with the shape expected from prompt gamma rays in the clinical environment, and the flux is also bunched with the accelerator frequency. The charge sharing effect of the CZT detector is studied qualitatively for different energy ranges. The LSO detector pixel discrimination resolution is analyzed and it shows a trend to improve for high energy depositions. The time correlation between the pulsed prompt photons and the measured detector signals, to be used for background suppression, exhibits a time resolution of 3 ns FWHM for the CZT detector and of 2 ns for the LSO detector. A time walk correction and pixel-wise calibration is applied for the LSO detector, whose resolution improves up to 630 ps. In conclusion, the detector setup is suitable for time-resolved background suppression in pulsed clinical particle accelerators. Ongoing tasks are the quantitative comparison with simulations and the test of imaging algorithms. Experiments at proton accelerators have also been performed and are currently under analysis.

  6. Analysis of pure tar substances (polycyclic aromatic hydrocarbons) in the gas stream using ultraviolet visible (UV-Vis) spectroscopy and multivariate curve resolution (MCR).

    PubMed

    Weide, Tobias; Guschin, Viktor; Becker, Wolfgang; Koelle, Sabine; Maier, Simon; Seidelt, Stephan

    2015-01-01

    The analysis of tar, mostly characterized as polycyclic aromatic hydrocarbons (PAHs), describes a topic that has been researched for years. An online analysis of tar in the gas stream in particular is needed to characterize the tar conversion or formation in the biomass gasification process. The online analysis in the gas is carried out with ultraviolet-visible (UV-Vis) spectroscopy (190-720 nm). This online analysis is performed with a measuring cell developed by the Fraunhofer Institute for Chemical Technology (ICT). To this day, online tar measurements using UV-Vis spectroscopy have not been carried out in detail. Therefore, PAHs are analyzed as follows. The measurements are split into different steps. The first step to prove the online method is to vaporize single tar substances. These experiments show that a qualitative analysis of PAHs in the gas stream with the used measurement setup is possible. Furthermore, it is shown that the method provides very exact results, so that a differentiation of various PAHs is possible. The next step is to vaporize a PAH mixture. This step consists of vaporizing five pure substances almost simultaneously. The interpretation of the resulting data is made using a chemometric interpretation method, the multivariate curve resolution (MCR). The verification of the calculated results is the main aim of this experiment. It has been shown that the tar mixture can be analyzed qualitatively and quantitatively (in arbitrary units) in detail using the MCR. Finally it is the main goal of this paper to show the first steps in the applicability of the UV-Vis spectroscopy and the measurement setup on online tar analysis in view of characterizing the biomass gasification process. Due to that, the gasification plant (at the laboratory scale), developed and constructed by the Fraunhofer ICT, has been used to vaporize these substances. Using this gasification plant for the experiments enables the usage of the measurement setup also for the spectroscopic analysis of the tar formation during the biomass gasification.

  7. SU-E-T-292: Sensitivity of Fractionated Lung IMPT Treatments to Setup Uncertainties and Motion Effects

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Dowdell, S; Grassberger, C; Paganetti, H

    2014-06-01

    Purpose: Evaluate the sensitivity of intensity-modulated proton therapy (IMPT) lung treatments to systematic and random setup uncertainties combined with motion effects. Methods: Treatment plans with single-field homogeneity restricted to ±20% (IMPT-20%) were compared to plans with no restriction (IMPT-full). 4D Monte Carlo simulations were performed for 10 lung patients using the patient CT geometry with either ±5mm systematic or random setup uncertainties applied over a 35 × 2.5Gy(RBE) fractionated treatment course. Intra-fraction, inter-field and inter-fraction motions were investigated. 50 fractionated treatments with systematic or random setup uncertainties applied to each fraction were generated for both IMPT delivery methods and threemore » energy-dependent spot sizes (big spots - BS σ=18-9mm, intermediate spots - IS σ=11-5mm, small spots - SS σ=4-2mm). These results were compared to a Monte Carlo recalculation of the original treatment plan, with results presented as the difference in EUD (ΔEUD), V{sub 95} (ΔV{sub 95}) and target homogeneity (ΔD{sub 1}–D{sub 99}) between the 4D simulations and the Monte Carlo calculation on the planning CT. Results: The standard deviations in the ΔEUD were 1.95±0.47(BS), 1.85±0.66(IS) and 1.31±0.35(SS) times higher in IMPT-full compared to IMPT-20% when ±5mm systematic setup uncertainties were applied. The ΔV{sub 95} variations were also 1.53±0.26(BS), 1.60±0.50(IS) and 1.38±0.38(SS) times higher for IMPT-full. For random setup uncertainties, the standard deviations of the ΔEUD from 50 simulated fractionated treatments were 1.94±0.90(BS), 2.13±1.08(IS) and 1.45±0.57(SS) times higher in IMPTfull compared to IMPT-20%. For all spot sizes considered, the ΔD{sub 1}-D{sub 99} coincided within the uncertainty limits for the two IMPT delivery methods, with the mean value always higher for IMPT-full. Statistical analysis showed significant differences between the IMPT-full and IMPT-20% dose distributions for the majority of scenarios studied. Conclusion: Lung IMPT-full treatments are more sensitive to both systematic and random setup uncertainties compared to IMPT-20%. This work was supported by the NIH R01 CA111590.« less

  8. A review of the design and clinical evaluation of the ShefStim array-based functional electrical stimulation system.

    PubMed

    Kenney, Laurence P; Heller, Ben W; Barker, Anthony T; Reeves, Mark L; Healey, Jamie; Good, Timothy R; Cooper, Glen; Sha, Ning; Prenton, Sarah; Liu, Anmin; Howard, David

    2016-11-01

    Functional electrical stimulation has been shown to be a safe and effective means of correcting foot drop of central neurological origin. Current surface-based devices typically consist of a single channel stimulator, a sensor for determining gait phase and a cuff, within which is housed the anode and cathode. The cuff-mounted electrode design reduces the likelihood of large errors in electrode placement, but the user is still fully responsible for selecting the correct stimulation level each time the system is donned. Researchers have investigated different approaches to automating aspects of setup and/or use, including recent promising work based on iterative learning techniques. This paper reports on the design and clinical evaluation of an electrode array-based FES system for the correction of drop foot, ShefStim. The paper reviews the design process from proof of concept lab-based study, through modelling of the array geometry and interface layer to array search algorithm development. Finally, the paper summarises two clinical studies involving patients with drop foot. The results suggest that the ShefStim system with automated setup produces results which are comparable with clinician setup of conventional systems. Further, the final study demonstrated that patients can use the system without clinical supervision. When used unsupervised, setup time was 14min (9min for automated search plus 5min for donning the equipment), although this figure could be reduced significantly with relatively minor changes to the design. Copyright © 2016 IPEM. Published by Elsevier Ltd. All rights reserved.

  9. Stochastic formulation of patient positioning using linac-mounted cone beam imaging with prior knowledge

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hoegele, W.; Loeschel, R.; Dobler, B.

    2011-02-15

    Purpose: In this work, a novel stochastic framework for patient positioning based on linac-mounted CB projections is introduced. Based on this formulation, the most probable shifts and rotations of the patient are estimated, incorporating interfractional deformations of patient anatomy and other uncertainties associated with patient setup. Methods: The target position is assumed to be defined by and is stochastically determined from positions of various features such as anatomical landmarks or markers in CB projections, i.e., radiographs acquired with a CB-CT system. The patient positioning problem of finding the target location from CB projections is posed as an inverse problem withmore » prior knowledge and is solved using a Bayesian maximum a posteriori (MAP) approach. The prior knowledge is three-fold and includes the accuracy of an initial patient setup (such as in-room laser and skin marks), the plasticity of the body (relative shifts between target and features), and the feature detection error in CB projections (which may vary depending on specific detection algorithm and feature type). For this purpose, MAP estimators are derived and a procedure of using them in clinical practice is outlined. Furthermore, a rule of thumb is theoretically derived, relating basic parameters of the prior knowledge (initial setup accuracy, plasticity of the body, and number of features) and the parameters of CB data acquisition (number of projections and accuracy of feature detection) to the expected estimation accuracy. Results: MAP estimation can be applied to arbitrary features and detection algorithms. However, to experimentally demonstrate its applicability and to perform the validation of the algorithm, a water-equivalent, deformable phantom with features represented by six 1 mm chrome balls were utilized. These features were detected in the cone beam projections (XVI, Elekta Synergy) by a local threshold method for demonstration purposes only. The accuracy of estimation (strongly varying for different plasticity parameters of the body) agreed with the rule of thumb formula. Moreover, based on this rule of thumb formula, about 20 projections for 6 detectable features seem to be sufficient for a target estimation accuracy of 0.2 cm, even for relatively large feature detection errors with standard deviation of 0.5 cm and spatial displacements of the features with standard deviation of 0.5 cm. Conclusions: The authors have introduced a general MAP-based patient setup algorithm accounting for different sources of uncertainties, which are utilized as the prior knowledge in a transparent way. This new framework can be further utilized for different clinical sites, as well as theoretical developments in the field of patient positioning for radiotherapy.« less

  10. An improved authenticated key agreement protocol for telecare medicine information system.

    PubMed

    Liu, Wenhao; Xie, Qi; Wang, Shengbao; Hu, Bin

    2016-01-01

    In telecare medicine information systems (TMIS), identity authentication of patients plays an important role and has been widely studied in the research field. Generally, it is realized by an authenticated key agreement protocol, and many such protocols were proposed in the literature. Recently, Zhang et al. pointed out that Islam et al.'s protocol suffers from the following security weaknesses: (1) Any legal but malicious patient can reveal other user's identity; (2) An attacker can launch off-line password guessing attack and the impersonation attack if the patient's identity is compromised. Zhang et al. also proposed an improved authenticated key agreement scheme with privacy protection for TMIS. However, in this paper, we point out that Zhang et al.'s scheme cannot resist off-line password guessing attack, and it fails to provide the revocation of lost/stolen smartcard. In order to overcome these weaknesses, we propose an improved protocol, the security and authentication of which can be proven using applied pi calculus based formal verification tool ProVerif.

  11. Parotid Glands Dose–Effect Relationships Based on Their Actually Delivered Doses: Implications for Adaptive Replanning in Radiation Therapy of Head-and-Neck Cancer

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hunter, Klaudia U.; Fernandes, Laura L.; Vineberg, Karen A.

    2013-11-15

    Purpose: Doses actually delivered to the parotid glands during radiation therapy often exceed planned doses. We hypothesized that the delivered doses correlate better with parotid salivary output than the planned doses, used in all previous studies, and that determining these correlations will help make decisions regarding adaptive radiation therapy (ART) aimed at reducing the delivered doses. Methods and Materials: In this prospective study, oropharyngeal cancer patients treated definitively with chemoirradiation underwent daily cone-beam computed tomography (CBCT) with clinical setup alignment based on the C2 posterior edge. Parotid glands in the CBCTs were aligned by deformable registration to calculate cumulative deliveredmore » doses. Stimulated salivary flow rates were measured separately from each parotid gland pretherapy and periodically posttherapy. Results: Thirty-six parotid glands of 18 patients were analyzed. Average mean planned doses was 32 Gy, and differences from planned to delivered mean gland doses were −4.9 to +8.4 Gy, median difference +2.2 Gy in glands in which delivered doses increased relative to planned. Both planned and delivered mean doses were significantly correlated with posttreatment salivary outputs at almost all posttherapy time points, without statistically significant differences in the correlations. Large dispersions (on average, SD 3.6 Gy) characterized the dose–effect relationships for both. The differences between the cumulative delivered doses and planned doses were evident at first fraction (r=.92, P<.0001) because of complex setup deviations (eg, rotations and neck articulations), uncorrected by the translational clinical alignments. Conclusions: After daily translational setup corrections, differences between planned and delivered doses in most glands were small relative to the SDs of the dose–saliva data, suggesting that ART is not likely to gain measurable salivary output improvement in most cases. These differences were observed at first treatment, indicating potential benefit for more complex setup corrections or adaptive interventions in the minority of patients with large deviations detected early by CBCT.« less

  12. TH-EF-BRB-11: Volumetric Modulated Arc Therapy for Total Body Irradiation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ouyang, L; Folkerts, M; Hrycushko, B

    Purpose: To develop a modern, patient-comfortable total body irradiation (TBI) technique suitable for standard-sized linac vaults. Methods: An indexed rotatable immobilization system (IRIS) was developed to make possible total-body CT imaging and radiation delivery on conventional couches. Treatment consists of multi-isocentric volumetric modulated arc therapy (VMAT) to the upper body and parallel-opposed fields to the lower body. Each isocenter is indexed to the couch and includes a 180° IRIS rotation between the upper and lower body fields. VMAT fields are optimized to satisfy lung dose objectives while achieving a uniform therapeutic dose to the torso. End-to-end tests with a randomore » phantom were used to verify dosimetric characteristics. Treatment plan robustness regarding setup uncertainty was assessed by simulating global and regional isocenter setup shifts on patient data sets. Dosimetric comparisons were made with conventional extended distance, standing TBI (cTBI) plans using a Monte Carlo-based calculation. Treatment efficiency was assessed for eight courses of patient treatment. Results: The IRIS system is level and orthogonal to the scanned CT image plane, with lateral shifts <2mm following rotation. End-to-end tests showed surface doses within ±10% of the prescription dose, field junction doses within ±15% of prescription dose. Plan robustness tests showed <15% changes in dose with global setup errors up to 5mm in each direction. Local 5mm relative setup errors in the chest resulted in < 5% dose changes. Local 5mm shift errors in the pelvic and upper leg junction resulted in <10% dose changes while a 10mm shift error causes dose changes up to 25%. Dosimetric comparison with cTBI showed VMAT-TBI has advantages in preserving chest wall dose with flexibility in leveraging the PTV-body and PTV-lung dose. Conclusion: VMAT-TBI with the IRIS system was shown clinically feasible as a cost-effective approach to TBI for standard-sized linac vaults.« less

  13. A calibration method for patient specific IMRT QA using a single therapy verification film

    PubMed Central

    Shukla, Arvind Kumar; Oinam, Arun S.; Kumar, Sanjeev; Sandhu, I.S.; Sharma, S.C.

    2013-01-01

    Aim The aim of the present study is to develop and verify the single film calibration procedure used in intensity-modulated radiation therapy (IMRT) quality assurance. Background Radiographic films have been regularly used in routine commissioning of treatment modalities and verification of treatment planning system (TPS). The radiation dosimetery based on radiographic films has ability to give absolute two-dimension dose distribution and prefer for the IMRT quality assurance. However, the single therapy verification film gives a quick and significant reliable method for IMRT verification. Materials and methods A single extended dose rate (EDR 2) film was used to generate the sensitometric curve of film optical density and radiation dose. EDR 2 film was exposed with nine 6 cm × 6 cm fields of 6 MV photon beam obtained from a medical linear accelerator at 5-cm depth in solid water phantom. The nine regions of single film were exposed with radiation doses raging from 10 to 362 cGy. The actual dose measurements inside the field regions were performed using 0.6 cm3 ionization chamber. The exposed film was processed after irradiation using a VIDAR film scanner and the value of optical density was noted for each region. Ten IMRT plans of head and neck carcinoma were used for verification using a dynamic IMRT technique, and evaluated using the gamma index method with TPS calculated dose distribution. Results Sensitometric curve has been generated using a single film exposed at nine field region to check quantitative dose verifications of IMRT treatments. The radiation scattered factor was observed to decrease exponentially with the increase in the distance from the centre of each field region. The IMRT plans based on calibration curve were verified using the gamma index method and found to be within acceptable criteria. Conclusion The single film method proved to be superior to the traditional calibration method and produce fast daily film calibration for highly accurate IMRT verification. PMID:24416558

  14. SU-F-T-269: Preliminary Experience of Kuwait Cancer Control Center (KCCC) On IMRT Treatment Planning and Pre-Treatment Verification

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sethuraman, TKR; Sherif, M; Subramanian, N

    Purpose: The complexity of IMRT delivery requires pre-treatment quality assurance and plan verification. KCCC has implemented IMRT clinically in few sites and will extend to all sites. Recently, our Varian linear accelerator and Eclipse planning system were upgraded from Millennium 80 to 120 Multileaf Collimator (MLC) and from v8.6 to 11.0 respectively. Our preliminary experience on the pre-treatment quality assurance verification is discussed. Methods: Eight Breast, Three Prostate and One Hypopharynx cancer patients were planned with step and shoot IMRT. All breast cases were planned before the upgrade with 60% cases treated. The ICRU 83 recommendations were followed for themore » dose prescription and constraints to OAR for all cases. Point dose measurement was done with CIRS cylindrical phantom and PTW 0.125 cc ionization chamber. Measured dose was compared with calculated dose at the point of measurement. Map CHECK diode array phantom was used for the plan verification. Planned and measured doses were compared by applying gamma index of 3% (dose difference) / 3 mm DTA (average distance to agreement). For all cases, a plan is considered to be successful if more than 95% of the tested diodes pass the gamma test. A prostate case was chosen to compare the plan verification before and after the upgrade. Results: Point dose measurement results were in agreement with the calculated doses. The maximum deviation observed was 2.3%. The passing rate of average gamma index was measured higher than 97% for the plan verification of all cases. Similar result was observed for plan verification of the chosen prostate case before and after the upgrade. Conclusion: Our preliminary experience from the obtained results validates the accuracy of our QA process and provides confidence to extend IMRT to all sites in Kuwait.« less

  15. Characterization of a phantom setup for breast conserving cancer surgery

    NASA Astrophysics Data System (ADS)

    Chadwell, Jacob T.; Conley, Rebekah H.; Collins, Jarrod A.; Meszoely, Ingrid M.; Miga, Michael I.

    2016-03-01

    The purpose of this work is to develop an anatomically and mechanically representative breast phantom for the validation of breast conserving surgical therapies, specifically, in this case, image guided surgeries. Using three patients scheduled for lumpectomy and four healthy volunteers in mock surgical presentations, the magnitude, direction, and location of breast deformations was analyzed. A phantom setup was then designed to approximate such deformations in a mock surgical environment. Specifically, commercially available and custom-built polyvinyl alcohol (PVA) phantoms were used to mimic breast tissue during surgery. A custom designed deformation apparatus was then created to reproduce deformations seen in typical clinical setups of the pre- and intra-operative breast geometry. Quantitative analysis of the human subjects yielded a positive correlation between breast volume and amount of breast deformation. Phantom results reflected similar behavior with the custom-built PVA phantom outperforming the commercial phantom.

  16. Metal artifact suppression in megavoltage computed tomography

    NASA Astrophysics Data System (ADS)

    Schreiner, L. John; Rogers, Myron; Salomons, Greg; Kerr, Andrew

    2005-04-01

    There has been considerable interest in megavoltage CT (MVCT) imaging associated with the development of image guided radiation therapy. It is clear that MVCT can provide good image quality for patient setup verification with soft tissue contrast much better than noted in conventional megavoltage portal imaging. In addition, it has been observed that MVCT images exhibit considerably reduced artifacts surrounding metal implants (e.g., surgical clips, hip implants, dental fillings) compared to conventional diagnostic CT images (kVCT). When encountered, these artifacts greatly limit the usefulness of kVCT images, and a variety of solutions have been proposed to remove the artifacts, but these have met with only partial success. In this paper, we investigate the potential for CT imaging in regions surrounding metal implants using high-energy photons from a Cobalt-60 source and from a 4 MV linear accelerator. MVCT and kVCT images of contrast phantoms and a phantom containing a hip prosthesis are compared and analysed. We show that MVCT scans provide good fidelity for CT number quantification in the high-density regions of the images, and in the regions immediately adjacent to the metal implants. They also provide structural details within the high-density inserts and implants. Calculations will show that practical clinical MVCT imaging, able to detect 3% contrast objects, should be achievable with doses of about 2.5cGy. This suggests that MVCT not only has a role in radiotherapy treatment planning and guidance, but may also be indicated for surgical guidance and follow-up in regions where metal implants cannot be avoided.

  17. Differentiation of tissue and kidney stones for laser lithotripsy using different spectroscopic approaches

    NASA Astrophysics Data System (ADS)

    Lange, Birgit; Cordes, Jens; Brinkmann, Ralf

    2015-07-01

    Holmium lasers are nowadays the gold standard for endoscopic laser lithotripsy. However, there is a risk of damaging or perforating the ureter or kidney tissue when the vision is poor. An automatic tissue/stone differentiation would improve the handling and safety of the procedure. To achieve this objective, an easy and robust real-time discrimination method has to be found which can be used to realize a feedback loop to control the laser system. Two possible approaches have been evaluated: White light reflectance and fluorescence spectroscopy. In both cases, we use the treatment fiber for detection and evaluate the possibility to decide whether the fiber is placed in front of tissue or calculus by the signal that is delivered by the surface in front of it. White light reflectance spectroscopy uses the standard light source for endourologic surgeries: Radiation of a Xenon light source is coupled to the ureteroscope via a liquid light guide. The part of the white light that is reflected back into the fiber is spectroscopically analyzed. In a clinical proof of concept study reflection signals were measured in vivo in 8 patients. For differentiation of stone and tissue via autofluorescence, excitation as well as detection was done via the treatment fiber. A suitable excitation wavelength was chosen with in vitro measurements (UV / visible) on several human renal calculi and porcine tissues. For verification of the positive results with green excitation in a clinical proof of concept study, a measurement set-up was realized which allows the recording of fluorescence signals during an endourological intervention.

  18. Biomarker Discovery and Verification of Esophageal Squamous Cell Carcinoma Using Integration of SWATH/MRM.

    PubMed

    Hou, Guixue; Lou, Xiaomin; Sun, Yulin; Xu, Shaohang; Zi, Jin; Wang, Quanhui; Zhou, Baojin; Han, Bo; Wu, Lin; Zhao, Xiaohang; Lin, Liang; Liu, Siqi

    2015-09-04

    We propose an efficient integration of SWATH with MRM for biomarker discovery and verification when the corresponding ion library is well established. We strictly controlled the false positive rate associated with SWATH MS signals and carefully selected the target peptides coupled with SWATH and MRM. We collected 10 samples of esophageal squamous cell carcinoma (ESCC) tissues paired with tumors and adjacent regions and quantified 1758 unique proteins with FDR 1% at protein level using SWATH, in which 467 proteins were abundance-dependent with ESCC. After carefully evaluating the SWATH MS signals of the up-regulated proteins, we selected 120 proteins for MRM verification. MRM analysis of the pooled and individual esophageal tissues resulted in 116 proteins that exhibited similar abundance response modes to ESCC that were acquired with SWATH. Because the ESCC-related proteins consisted of a high percentile of secreted proteins, we conducted the MRM assay on patient sera that were collected from pre- and postoperation. Of the 116 target proteins, 42 were identified in the ESCC sera, including 11 with lowered abundances postoperation. Coupling SWATH and MRM is thus feasible and efficient for the discovery and verification of cancer-related protein biomarkers.

  19. Intersubject variability and intrasubject reproducibility of 12-lead ECG metrics: Implications for human verification.

    PubMed

    Jekova, Irena; Krasteva, Vessela; Leber, Remo; Schmid, Ramun; Twerenbold, Raphael; Müller, Christian; Reichlin, Tobias; Abächerli, Roger

    Electrocardiogram (ECG) biometrics is an advanced technology, not yet covered by guidelines on criteria, features and leads for maximal authentication accuracy. This study aims to define the minimal set of morphological metrics in 12-lead ECG by optimization towards high reliability and security, and validation in a person verification model across a large population. A standard 12-lead resting ECG database from 574 non-cardiac patients with two remote recordings (>1year apart) was used. A commercial ECG analysis module (Schiller AG) measured 202 morphological features, including lead-specific amplitudes, durations, ST-metrics, and axes. Coefficient of variation (CV, intersubject variability) and percent-mean-absolute-difference (PMAD, intrasubject reproducibility) defined the optimization (PMAD/CV→min) and restriction (CV<30%) criteria for selection of the most stable and distinctive features. Linear discriminant analysis (LDA) validated the non-redundant feature set for person verification. Maximal LDA verification sensitivity (85.3%) and specificity (86.4%) were validated for 11 optimal features: R-amplitude (I,II,V1,V2,V3,V5), S-amplitude (V1,V2), Tnegative-amplitude (aVR), and R-duration (aVF,V1). Copyright © 2016 Elsevier Inc. All rights reserved.

  20. An Analysis of Plan Robustness for Esophageal Tumors: Comparing Volumetric Modulated Arc Therapy Plans and Spot Scanning Proton Planning

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Warren, Samantha, E-mail: samantha.warren@oncology.ox.ac.uk; Partridge, Mike; Bolsi, Alessandra

    Purpose: Planning studies to compare x-ray and proton techniques and to select the most suitable technique for each patient have been hampered by the nonequivalence of several aspects of treatment planning and delivery. A fair comparison should compare similarly advanced delivery techniques from current clinical practice and also assess the robustness of each technique. The present study therefore compared volumetric modulated arc therapy (VMAT) and single-field optimization (SFO) spot scanning proton therapy plans created using a simultaneous integrated boost (SIB) for dose escalation in midesophageal cancer and analyzed the effect of setup and range uncertainties on these plans. Methods andmore » Materials: For 21 patients, SIB plans with a physical dose prescription of 2 Gy or 2.5 Gy/fraction in 25 fractions to planning target volume (PTV){sub 50Gy} or PTV{sub 62.5Gy} (primary tumor with 0.5 cm margins) were created and evaluated for robustness to random setup errors and proton range errors. Dose–volume metrics were compared for the optimal and uncertainty plans, with P<.05 (Wilcoxon) considered significant. Results: SFO reduced the mean lung dose by 51.4% (range 35.1%-76.1%) and the mean heart dose by 40.9% (range 15.0%-57.4%) compared with VMAT. Proton plan robustness to a 3.5% range error was acceptable. For all patients, the clinical target volume D{sub 98} was 95.0% to 100.4% of the prescribed dose and gross tumor volume (GTV) D{sub 98} was 98.8% to 101%. Setup error robustness was patient anatomy dependent, and the potential minimum dose per fraction was always lower with SFO than with VMAT. The clinical target volume D{sub 98} was lower by 0.6% to 7.8% of the prescribed dose, and the GTV D{sub 98} was lower by 0.3% to 2.2% of the prescribed GTV dose. Conclusions: The SFO plans achieved significant sparing of normal tissue compared with the VMAT plans for midesophageal cancer. The target dose coverage in the SIB proton plans was less robust to random setup errors and might be unacceptable for certain patients. Robust optimization to ensure adequate target coverage of SIB proton plans might be beneficial.« less

  1. An Image-Guided Study of Setup Reproducibility of Postmastectomy Breast Cancer Patients Treated With Inverse-Planned Intensity Modulated Radiation Therapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Feng, Christine H.; Gerry, Emily; Chmura, Steven J.

    2015-01-01

    Purpose: To calculate planning target volume (PTV) margins for chest wall and regional nodal targets using daily orthogonal kilovolt (kV) imaging and to study residual setup error after kV alignment using volumetric cone-beam computed tomography (CBCT). Methods and Materials: Twenty-one postmastectomy patients were treated with intensity modulated radiation therapy with 7-mm PTV margins. Population-based PTV margins were calculated from translational shifts after daily kV positioning and/or weekly CBCT data for each of 8 patients, whose surgical clips were used as surrogates for target volumes. Errors from kV and CBCT data were mathematically combined to generate PTV margins for 3 simulatedmore » alignment workflows: (1) skin marks alone; (2) weekly kV imaging; and (3) daily kV imaging. Results: The kV data from 613 treatment fractions indicated that a 7-mm uniform margin would account for 95% of daily shifts if patients were positioned using only skin marks. Total setup errors incorporating both kV and CBCT data were larger than those from kV alone, yielding PTV expansions of 7 mm anterior–posterior, 9 mm left–right, and 9 mm superior–inferior. Required PTV margins after weekly kV imaging were similar in magnitude as alignment to skin marks, but rotational adjustments of patients were required in 32% ± 17% of treatments. These rotations would have remained uncorrected without the use of daily kV imaging. Despite the use of daily kV imaging, CBCT data taken at the treatment position indicate that an anisotropic PTV margin of 6 mm anterior–posterior, 4 mm left–right, and 8 mm superior–inferior must be retained to account for residual errors. Conclusions: Cone-beam CT provides additional information on 3-dimensional reproducibility of treatment setup for chest wall targets. Three-dimensional data indicate that a uniform 7-mm PTV margin is insufficient in the absence of daily IGRT. Interfraction movement is greater than suggested by 2-dimensional imaging, thus a margin of at least 4 to 8 mm must be retained despite the use of daily IGRT.« less

  2. An Analysis of Plan Robustness for Esophageal Tumors: Comparing Volumetric Modulated Arc Therapy Plans and Spot Scanning Proton Planning

    PubMed Central

    Warren, Samantha; Partridge, Mike; Bolsi, Alessandra; Lomax, Anthony J.; Hurt, Chris; Crosby, Thomas; Hawkins, Maria A.

    2016-01-01

    Purpose Planning studies to compare x-ray and proton techniques and to select the most suitable technique for each patient have been hampered by the nonequivalence of several aspects of treatment planning and delivery. A fair comparison should compare similarly advanced delivery techniques from current clinical practice and also assess the robustness of each technique. The present study therefore compared volumetric modulated arc therapy (VMAT) and single-field optimization (SFO) spot scanning proton therapy plans created using a simultaneous integrated boost (SIB) for dose escalation in midesophageal cancer and analyzed the effect of setup and range uncertainties on these plans. Methods and Materials For 21 patients, SIB plans with a physical dose prescription of 2 Gy or 2.5 Gy/fraction in 25 fractions to planning target volume (PTV)50Gy or PTV62.5Gy (primary tumor with 0.5 cm margins) were created and evaluated for robustness to random setup errors and proton range errors. Dose–volume metrics were compared for the optimal and uncertainty plans, with P<.05 (Wilcoxon) considered significant. Results SFO reduced the mean lung dose by 51.4% (range 35.1%-76.1%) and the mean heart dose by 40.9% (range 15.0%-57.4%) compared with VMAT. Proton plan robustness to a 3.5% range error was acceptable. For all patients, the clinical target volume D98 was 95.0% to 100.4% of the prescribed dose and gross tumor volume (GTV) D98 was 98.8% to 101%. Setup error robustness was patient anatomy dependent, and the potential minimum dose per fraction was always lower with SFO than with VMAT. The clinical target volume D98 was lower by 0.6% to 7.8% of the prescribed dose, and the GTV D98 was lower by 0.3% to 2.2% of the prescribed GTV dose. Conclusions The SFO plans achieved significant sparing of normal tissue compared with the VMAT plans for midesophageal cancer. The target dose coverage in the SIB proton plans was less robust to random setup errors and might be unacceptable for certain patients. Robust optimization to ensure adequate target coverage of SIB proton plans might be beneficial. PMID:27084641

  3. An Analysis of Plan Robustness for Esophageal Tumors: Comparing Volumetric Modulated Arc Therapy Plans and Spot Scanning Proton Planning.

    PubMed

    Warren, Samantha; Partridge, Mike; Bolsi, Alessandra; Lomax, Anthony J; Hurt, Chris; Crosby, Thomas; Hawkins, Maria A

    2016-05-01

    Planning studies to compare x-ray and proton techniques and to select the most suitable technique for each patient have been hampered by the nonequivalence of several aspects of treatment planning and delivery. A fair comparison should compare similarly advanced delivery techniques from current clinical practice and also assess the robustness of each technique. The present study therefore compared volumetric modulated arc therapy (VMAT) and single-field optimization (SFO) spot scanning proton therapy plans created using a simultaneous integrated boost (SIB) for dose escalation in midesophageal cancer and analyzed the effect of setup and range uncertainties on these plans. For 21 patients, SIB plans with a physical dose prescription of 2 Gy or 2.5 Gy/fraction in 25 fractions to planning target volume (PTV)50Gy or PTV62.5Gy (primary tumor with 0.5 cm margins) were created and evaluated for robustness to random setup errors and proton range errors. Dose-volume metrics were compared for the optimal and uncertainty plans, with P<.05 (Wilcoxon) considered significant. SFO reduced the mean lung dose by 51.4% (range 35.1%-76.1%) and the mean heart dose by 40.9% (range 15.0%-57.4%) compared with VMAT. Proton plan robustness to a 3.5% range error was acceptable. For all patients, the clinical target volume D98 was 95.0% to 100.4% of the prescribed dose and gross tumor volume (GTV) D98 was 98.8% to 101%. Setup error robustness was patient anatomy dependent, and the potential minimum dose per fraction was always lower with SFO than with VMAT. The clinical target volume D98 was lower by 0.6% to 7.8% of the prescribed dose, and the GTV D98 was lower by 0.3% to 2.2% of the prescribed GTV dose. The SFO plans achieved significant sparing of normal tissue compared with the VMAT plans for midesophageal cancer. The target dose coverage in the SIB proton plans was less robust to random setup errors and might be unacceptable for certain patients. Robust optimization to ensure adequate target coverage of SIB proton plans might be beneficial. Copyright © 2016 The Authors. Published by Elsevier Inc. All rights reserved.

  4. A uniqueness-and-anonymity-preserving remote user authentication scheme for connected health care.

    PubMed

    Chang, Ya-Fen; Yu, Shih-Hui; Shiao, Ding-Rui

    2013-04-01

    Connected health care provides new opportunities for improving financial and clinical performance. Many connected health care applications such as telecare medicine information system, personally controlled health records system, and patient monitoring have been proposed. Correct and quality care is the goal of connected heath care, and user authentication can ensure the legality of patients. After reviewing authentication schemes for connected health care applications, we find that many of them cannot protect patient privacy such that others can trace users/patients by the transmitted data. And the verification tokens used by these authentication schemes to authenticate users or servers are only password, smart card and RFID tag. Actually, these verification tokens are not unique and easy to copy. On the other hand, biometric characteristics, such as iris, face, voiceprint, fingerprint and so on, are unique, easy to be verified, and hard to be copied. In this paper, a biometrics-based user authentication scheme will be proposed to ensure uniqueness and anonymity at the same time. With the proposed scheme, only the legal user/patient himself/herself can access the remote server, and no one can trace him/her according to transmitted data.

  5. SU-D-201-05: On the Automatic Recognition of Patient Safety Hazards in a Radiotherapy Setup Using a Novel 3D Camera System and a Deep Learning Framework

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Santhanam, A; Min, Y; Beron, P

    Purpose: Patient safety hazards such as a wrong patient/site getting treated can lead to catastrophic results. The purpose of this project is to automatically detect potential patient safety hazards during the radiotherapy setup and alert the therapist before the treatment is initiated. Methods: We employed a set of co-located and co-registered 3D cameras placed inside the treatment room. Each camera provided a point-cloud of fraxels (fragment pixels with 3D depth information). Each of the cameras were calibrated using a custom-built calibration target to provide 3D information with less than 2 mm error in the 500 mm neighborhood around the isocenter.more » To identify potential patient safety hazards, the treatment room components and the patient’s body needed to be identified and tracked in real-time. For feature recognition purposes, we used a graph-cut based feature recognition with principal component analysis (PCA) based feature-to-object correlation to segment the objects in real-time. Changes in the object’s position were tracked using the CamShift algorithm. The 3D object information was then stored for each classified object (e.g. gantry, couch). A deep learning framework was then used to analyze all the classified objects in both 2D and 3D and was then used to fine-tune a convolutional network for object recognition. The number of network layers were optimized to identify the tracked objects with >95% accuracy. Results: Our systematic analyses showed that, the system was effectively able to recognize wrong patient setups and wrong patient accessories. The combined usage of 2D camera information (color + depth) enabled a topology-preserving approach to verify patient safety hazards in an automatic manner and even in scenarios where the depth information is partially available. Conclusion: By utilizing the 3D cameras inside the treatment room and a deep learning based image classification, potential patient safety hazards can be effectively avoided.« less

  6. Impact of a quality-assessment dashboard on the comprehensive review of pharmacist performance.

    PubMed

    Trinh, Long D; Roach, Erin M; Vogan, Eric D; Lam, Simon W; Eggers, Garrett G

    2017-09-01

    The impact of a quality-assessment dashboard and individualized pharmacist performance feedback on the adherence of order verification was evaluated. A before-and-after study was conducted at a 1,440-bed academic medical center. Adherence of order verification was defined as orders verified according to institution-derived, medication-related guidelines and policies. Formulas were developed to assess the adherence of verified orders to dosing guidelines using patient-specific height, weight, and serum creatinine clearance values from the electronic medical record at the time of pharmacist verification. A total of 5 medications were assessed by the formulas for adherence and displayed on the dashboard: ampicillin-sulbactam, ciprofloxacin, piperacillin-tazobactam, acyclovir, and enoxaparin. Adherence of order verification was assessed before (May 1-July 31, 2015) and after (November 1, 2015-January 31, 2016) individualized performance feedback was given based on trends identified by the quality-assessment dashboard. There was a significant increase in the overall adherence rate postintervention (90.1% versus 91.9%, p = 0.040). Among the 34 pharmacists who participated, the percentage of pharmacists with at least 90% overall adherence increased postintervention (52.9% versus 70.6%, p = 0.103). Time to verification was similar before and after the study intervention (median, 6.0 minutes; interquartile range, 3-13 minutes). The rate of documentation for nonadherent orders increased significantly postintervention (57.1% versus 68.5%, p = 0.019). The implementation of the quality-assessment dashboard, educational sessions, and individualized performance feedback significantly improved pharmacist order-verification adherence to institution-derived, medication-related guidelines and policies and the documentation rate of nonadherent orders. Copyright © 2017 by the American Society of Health-System Pharmacists, Inc. All rights reserved.

  7. 4D offline PET-based treatment verification in scanned ion beam therapy: a phantom study

    NASA Astrophysics Data System (ADS)

    Kurz, Christopher; Bauer, Julia; Unholtz, Daniel; Richter, Daniel; Stützer, Kristin; Bert, Christoph; Parodi, Katia

    2015-08-01

    At the Heidelberg Ion-Beam Therapy Center, patient irradiation with scanned proton and carbon ion beams is verified by offline positron emission tomography (PET) imaging: the {β+} -activity measured within the patient is compared to a prediction calculated on the basis of the treatment planning data in order to identify potential delivery errors. Currently, this monitoring technique is limited to the treatment of static target structures. However, intra-fractional organ motion imposes considerable additional challenges to scanned ion beam radiotherapy. In this work, the feasibility and potential of time-resolved (4D) offline PET-based treatment verification with a commercial full-ring PET/CT (x-ray computed tomography) device are investigated for the first time, based on an experimental campaign with moving phantoms. Motion was monitored during the gated beam delivery as well as the subsequent PET acquisition and was taken into account in the corresponding 4D Monte-Carlo simulations and data evaluation. Under the given experimental conditions, millimeter agreement between the prediction and measurement was found. Dosimetric consequences due to the phantom motion could be reliably identified. The agreement between PET measurement and prediction in the presence of motion was found to be similar as in static reference measurements, thus demonstrating the potential of 4D PET-based treatment verification for future clinical applications.

  8. Is ExacTrac x-ray system an alternative to CBCT for positioning patients with head and neck cancers?

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Clemente, Stefania; Chiumento, Costanza; Fiorentino, Alba

    Purpose: To evaluate the usefulness of a six-degrees-of freedom (6D) correction using ExacTrac robotics system in patients with head-and-neck (HN) cancer receiving radiation therapy.Methods: Local setup accuracy was analyzed for 12 patients undergoing intensity-modulated radiation therapy (IMRT). Patient position was imaged daily upon two different protocols, cone-beam computed tomography (CBCT), and ExacTrac (ET) images correction. Setup data from either approach were compared in terms of both residual errors after correction and punctual displacement of selected regions of interest (Mandible, C2, and C6 vertebral bodies).Results: On average, both protocols achieved reasonably low residual errors after initial correction. The observed differences inmore » shift vectors between the two protocols showed that CBCT tends to weight more C2 and C6 at the expense of the mandible, while ET tends to average more differences among the different ROIs.Conclusions: CBCT, even without 6D correction capabilities, seems preferable to ET for better consistent alignment and the capability to see soft tissues. Therefore, in our experience, CBCT represents a benchmark for positioning head and neck cancer patients.« less

  9. Animal Robot Assisted-therapy for Rehabilitation of Patient with Post-Stroke Depression

    NASA Astrophysics Data System (ADS)

    Zikril Zulkifli, Winal; Shamsuddin, Syamimi; Hwee, Lim Thiam

    2017-06-01

    Recently, the utilization of therapeutic animal robots has expanded. This research aims to explore robotics application for mental healthcare in Malaysia through human-robot interaction (HRI). PARO, the robotic seal PARO was developed to give psychological effects on humans. Major Depressive Disorder (MDD) is a common but severe mood disorder. This study focuses on the interaction protocol between PARO and patients with MDD. Initially, twelve rehabilitation patients gave subjective evaluation on their first interaction with PARO. Next, therapeutic interaction environment was set-up with PARO in it to act as an augmentation strategy with other psychological interventions for post-stroke depression. Patient was exposed to PARO for 20 minutes. The results of behavioural analysis complemented with information from HRI survey question. The analysis also observed that the individual interactors engaged with the robot in diverse ways based on their needs Results show positive reaction toward the acceptance of an animal robot. Next, therapeutic interaction is set-up for PARO to contribute as an augmentation strategy with other psychological interventions for post-stroke depression. The outcome is to reduce the stress level among patients through facilitated therapy session with PARO

  10. PATIENT STUDY OF IN VIVO VERIFICATION OF BEAM DELIVERY AND RANGE, USING POSITRON EMISSION TOMOGRAPHY AND COMPUTED TOMOGRAPHY IMAGING AFTER PROTON THERAPY

    PubMed Central

    Parodi, Katia; Paganetti, Harald; Shih, Helen A.; Michaud, Susan; Loeffler, Jay S.; Delaney, Thomas F.; Liebsch, Norbert J.; Munzenrider, John E.; Fischman, Alan J.; Knopf, Antje; Bortfeld, Thomas

    2007-01-01

    Purpose To investigate the feasibility and value of positron emission tomography and computed tomography (PET/CT) for treatment verification after proton radiotherapy. Methods and Materials This study included 9 patients with tumors in the cranial base, spine, orbit, and eye. Total doses of 1.8–3 GyE and 10 GyE (for an ocular melanoma) per fraction were delivered in 1 or 2 fields. Imaging was performed with a commercial PET/CT scanner for 30 min, starting within 20 min after treatment. The same treatment immobilization device was used during imaging for all but 2 patients. Measured PET/CT images were coregistered to the planning CT and compared with the corresponding PET expectation, obtained from CT-based Monte Carlo calculations complemented by functional information. For the ocular case, treatment position was approximately replicated, and spatial correlation was deduced from reference clips visible in both the planning radiographs and imaging CT. Here, the expected PET image was obtained from an analytical model. Results Good spatial correlation and quantitative agreement within 30% were found between the measured and expected activity. For head-and-neck patients, the beam range could be verified with an accuracy of 1–2 mm in well-coregistered bony structures. Low spine and eye sites indicated the need for better fixation and coregistration methods. An analysis of activity decay revealed as tissue-effective half-lives of 800–1,150 s. Conclusions This study demonstrates the feasibility of postradiation PET/CT for in vivo treatment verification. It also indicates some technological and methodological improvements needed for optimal clinical application. PMID:17544003

  11. 42 CFR 493.1255 - Standard: Calibration and calibration verification procedures.

    Code of Federal Regulations, 2013 CFR

    2013-10-01

    ... affect the range used to report patient test results, and control values are not adversely affected by... that may influence test performance. (iii) Control materials reflect an unusual trend or shift, or are...

  12. 42 CFR 493.1255 - Standard: Calibration and calibration verification procedures.

    Code of Federal Regulations, 2011 CFR

    2011-10-01

    ... affect the range used to report patient test results, and control values are not adversely affected by... that may influence test performance. (iii) Control materials reflect an unusual trend or shift, or are...

  13. 42 CFR 493.1255 - Standard: Calibration and calibration verification procedures.

    Code of Federal Regulations, 2012 CFR

    2012-10-01

    ... affect the range used to report patient test results, and control values are not adversely affected by... that may influence test performance. (iii) Control materials reflect an unusual trend or shift, or are...

  14. 42 CFR 493.1255 - Standard: Calibration and calibration verification procedures.

    Code of Federal Regulations, 2014 CFR

    2014-10-01

    ... affect the range used to report patient test results, and control values are not adversely affected by... that may influence test performance. (iii) Control materials reflect an unusual trend or shift, or are...

  15. Multifractionated image-guided and stereotactic intensity-modulated radiotherapy of paraspinal tumors: a preliminary report.

    PubMed

    Yamada, Yoshiya; Lovelock, D Michael; Yenice, Kamil M; Bilsky, Mark H; Hunt, Margaret A; Zatcky, Joan; Leibel, Steven A

    2005-05-01

    The use of image-guided and stereotactic intensity-modulated radiotherapy (IMRT) techniques have made the delivery of high-dose radiation to lesions within close proximity to the spinal cord feasible. This report presents clinical and physical data regarding the use of IMRT coupled with noninvasive body frames (stereotactic and image-guided) for multifractionated radiotherapy. The Memorial Sloan-Kettering Cancer Center (Memorial) stereotactic body frame (MSBF) and Memorial body cradle (MBC) have been developed as noninvasive immobilizing devices for paraspinal IMRT using stereotactic (MSBF) and image-guided (MBC) techniques. Patients were either previously irradiated or prescribed doses beyond spinal cord tolerance (54 Gy in standard fractionation) and had unresectable gross disease involving the spinal canal. The planning target volume (PTV) was the gross tumor volume with a 1 cm margin. The PTV was not allowed to include the spinal cord contour. All treatment planning was performed using software developed within the institution. Isocenter verification was performed with an in-room computed tomography scan (MSBF) or electronic portal imaging devices, or both. Patients were followed up with serial magnetic resonance imaging every 3-4 months, and no patients were lost to follow-up. Kaplan-Meier statistics were used for analysis of clinical data. Both the MSBF and MBC were able to provide setup accuracy within 2 mm. With a median follow-up of 11 months, 35 patients (14 primary and 21 secondary malignancies) underwent treatment. The median dose previously received was 3000 cGy in 10 fractions. The median dose prescribed for these patients was 2000 cGy/5 fractions (2000-3000 cGy), which provided a median PTV V100 of 88%. In previously unirradiated patients, the median prescribed dose was 7000 cGy (5940-7000 cGy) with a median PTV V100 of 90%. The median Dmax to the cord was 34% and 68% for previously irradiated and never irradiated patients, respectively. More than 90% of patients experienced palliation from pain, weakness, or paresthesia; 75% and 81% of secondary and primary lesions, respectively, exhibited local control at the time of last follow-up. No cases of radiation-induced myelopathy or radiculopathy have thus far been encountered. Precision stereotactic and image-guided paraspinal IMRT allows the delivery of high doses of radiation in multiple fractions to tumors within close proximity to the spinal cord while respecting cord tolerance. Although preliminary, the clinical results are encouraging.

  16. Precision assessment of model-based RSA for a total knee prosthesis in a biplanar set-up.

    PubMed

    Trozzi, C; Kaptein, B L; Garling, E H; Shelyakova, T; Russo, A; Bragonzoni, L; Martelli, S

    2008-10-01

    Model-based Roentgen Stereophotogrammetric Analysis (RSA) was recently developed for the measurement of prosthesis micromotion. Its main advantage is that markers do not need to be attached to the implants as traditional marker-based RSA requires. Model-based RSA has only been tested in uniplanar radiographic set-ups. A biplanar set-up would theoretically facilitate the pose estimation algorithm, since radiographic projections would show more different shape features of the implants than in uniplanar images. We tested the precision of model-based RSA and compared it with that of the traditional marker-based method in a biplanar set-up. Micromotions of both tibial and femoral components were measured with both the techniques from double examinations of patients participating in a clinical study. The results showed that in the biplanar set-up model-based RSA presents a homogeneous distribution of precision for all the translation directions, but an inhomogeneous error for rotations, especially internal-external rotation presented higher errors than rotations about the transverse and sagittal axes. Model-based RSA was less precise than the marker-based method, although the differences were not significant for the translations and rotations of the tibial component, with the exception of the internal-external rotations. For both prosthesis components the precisions of model-based RSA were below 0.2 mm for all the translations, and below 0.3 degrees for rotations about transverse and sagittal axes. These values are still acceptable for clinical studies aimed at evaluating total knee prosthesis micromotion. In a biplanar set-up model-based RSA is a valid alternative to traditional marker-based RSA where marking of the prosthesis is an enormous disadvantage.

  17. Development of Biomarkers for Screening Hepatocellular Carcinoma Using Global Data Mining and Multiple Reaction Monitoring

    PubMed Central

    Yu, Su Jong; Jang, Eun Sun; Yu, Jiyoung; Cho, Geunhee; Yoon, Jung-Hwan; Kim, Youngsoo

    2013-01-01

    Hepatocellular carcinoma (HCC) is one of the most common and aggressive cancers and is associated with a poor survival rate. Clinically, the level of alpha-fetoprotein (AFP) has been used as a biomarker for the diagnosis of HCC. The discovery of useful biomarkers for HCC, focused solely on the proteome, has been difficult; thus, wide-ranging global data mining of genomic and proteomic databases from previous reports would be valuable in screening biomarker candidates. Further, multiple reaction monitoring (MRM), based on triple quadrupole mass spectrometry, has been effective with regard to high-throughput verification, complementing antibody-based verification pipelines. In this study, global data mining was performed using 5 types of HCC data to screen for candidate biomarker proteins: cDNA microarray, copy number variation, somatic mutation, epigenetic, and quantitative proteomics data. Next, we applied MRM to verify HCC candidate biomarkers in individual serum samples from 3 groups: a healthy control group, patients who have been diagnosed with HCC (Before HCC treatment group), and HCC patients who underwent locoregional therapy (After HCC treatment group). After determining the relative quantities of the candidate proteins by MRM, we compared their expression levels between the 3 groups, identifying 4 potential biomarkers: the actin-binding protein anillin (ANLN), filamin-B (FLNB), complementary C4-A (C4A), and AFP. The combination of 2 markers (ANLN, FLNB) improved the discrimination of the before HCC treatment group from the healthy control group compared with AFP. We conclude that the combination of global data mining and MRM verification enhances the screening and verification of potential HCC biomarkers. This efficacious integrative strategy is applicable to the development of markers for cancer and other diseases. PMID:23717429

  18. Development of biomarkers for screening hepatocellular carcinoma using global data mining and multiple reaction monitoring.

    PubMed

    Kim, Hyunsoo; Kim, Kyunggon; Yu, Su Jong; Jang, Eun Sun; Yu, Jiyoung; Cho, Geunhee; Yoon, Jung-Hwan; Kim, Youngsoo

    2013-01-01

    Hepatocellular carcinoma (HCC) is one of the most common and aggressive cancers and is associated with a poor survival rate. Clinically, the level of alpha-fetoprotein (AFP) has been used as a biomarker for the diagnosis of HCC. The discovery of useful biomarkers for HCC, focused solely on the proteome, has been difficult; thus, wide-ranging global data mining of genomic and proteomic databases from previous reports would be valuable in screening biomarker candidates. Further, multiple reaction monitoring (MRM), based on triple quadrupole mass spectrometry, has been effective with regard to high-throughput verification, complementing antibody-based verification pipelines. In this study, global data mining was performed using 5 types of HCC data to screen for candidate biomarker proteins: cDNA microarray, copy number variation, somatic mutation, epigenetic, and quantitative proteomics data. Next, we applied MRM to verify HCC candidate biomarkers in individual serum samples from 3 groups: a healthy control group, patients who have been diagnosed with HCC (Before HCC treatment group), and HCC patients who underwent locoregional therapy (After HCC treatment group). After determining the relative quantities of the candidate proteins by MRM, we compared their expression levels between the 3 groups, identifying 4 potential biomarkers: the actin-binding protein anillin (ANLN), filamin-B (FLNB), complementary C4-A (C4A), and AFP. The combination of 2 markers (ANLN, FLNB) improved the discrimination of the before HCC treatment group from the healthy control group compared with AFP. We conclude that the combination of global data mining and MRM verification enhances the screening and verification of potential HCC biomarkers. This efficacious integrative strategy is applicable to the development of markers for cancer and other diseases.

  19. Ramifications of the Children's Surgery Verification Program for Patients and Hospitals.

    PubMed

    Baxter, Katherine J; Gale, Bonnie F; Travers, Curtis D; Heiss, Kurt F; Raval, Mehul V

    2018-05-01

    The American College of Surgeons in 2015 instituted the Children's Surgery Verification program delineating requirements for hospitals providing pediatric surgical care. Our purpose was to examine possible effects of the Children's Surgery Verification program by evaluating neonates undergoing high-risk operations. Using the Kid's Inpatient Database 2009, we identified infants undergoing operations for 5 high-risk neonatal conditions. We considered all children's hospitals and children's units Level I centers and considered all others Level II/III. We estimated the number of neonates requiring relocation and the additional distance traveled. We used propensity score adjusted logistic regression to model mortality at Level I vs Level II/III hospitals. Overall, 7,938 neonates were identified across 21 states at 91 Level I and 459 Level II/III hospitals. Based on our classifications, 2,744 (34.6%) patients would need to relocate to Level I centers. The median additional distance traveled was 6.6 miles. The maximum distance traveled varied by state, from <55 miles (New Jersey and Rhode Island) to >200 miles (Montana, Oregon, Colorado, and California). The adjusted odds of mortality at Level II/III vs Level I centers was 1.67 (95% CI 1.44 to 1.93). We estimate 1 life would be saved for every 32 neonates moved. Although this conservative estimate demonstrates that more than one-third of complex surgical neonates in 2009 would have needed to relocate under the Children's Surgery Verification program, the additional distance traveled is relatively short for most but not all, and this program might improve mortality. Local level ramifications of this novel national program require additional investigation. Copyright © 2018 American College of Surgeons. Published by Elsevier Inc. All rights reserved.

  20. Implementation of a cloud-based electronic medical record for maternal and child health in rural Kenya.

    PubMed

    Haskew, John; Rø, Gunnar; Saito, Kaori; Turner, Kenrick; Odhiambo, George; Wamae, Annah; Sharif, Shahnaaz; Sugishita, Tomohiko

    2015-05-01

    Complete and timely health information is essential to inform public health decision-making for maternal and child health, but is often lacking in resource-constrained settings. Electronic medical record (EMR) systems are increasingly being adopted to support the delivery of health care, and are particularly amenable to maternal and child health services. An EMR system could enable the mother and child to be tracked and monitored throughout maternity shared care, improve quality and completeness of data collected and enhance sharing of health information between outpatient clinic and the hospital, and between clinical and public health services to inform decision-making. This study implemented a novel cloud-based electronic medical record system in a maternal and child health outpatient setting in Western Kenya between April and June 2013 and evaluated its impact on improving completeness of data collected by clinical and public health services. The impact of the system was assessed using a two-sample test of proportions pre- and post-implementation of EMR-based data verification. Significant improvements in completeness of the antenatal record were recorded through implementation of EMR-based data verification. A difference of 42.9% in missing data (including screening for hypertension, tuberculosis, malaria, HIV status or ART status of HIV positive women) was recorded pre- and post-implementation. Despite significant impact of EMR-based data verification on data completeness, overall screening rates in antenatal care were low. This study has shown that EMR-based data verification can improve the completeness of data collected in the patient record for maternal and child health. A number of issues, including data management and patient confidentiality, must be considered but significant improvements in data quality are recorded through implementation of this EMR model. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  1. SU-F-T-223: Radiotherapy Incident Reporting and Analysis System (RIRAS):Early Experience

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kapoor, R; Palta, J; Hagan, M

    Background & Purpose: RIRAS is a web-based information system deployed on the Veterans Health Administration intranet in early 2014 to collect adverse events and good catch data; analyze the causes and contributing factors; and find ways to prevent future occurrences. Material and Methods: Incident learning consists of a feedback loop which starts with reporting an event, followed by analysis of contributing factors, and culminates in the development of a patient safety work product (PSWP) to prevent recurrence. RIRAS permits both anonymous and non-anonymous reporting. Each report is analyzed by a team of medical physicists who are independent of the reportingmore » facility. The analysts usually contact the reporting facilities for additional information. We analyzed all reports and held telephonic interviews (when necessary) with the reporters. We then generated PSWPs with corrective/preventive and learning actions. Anonymous reporting is handled in the same manner, except without the ability to further interview the reporter. Results: In a significant number of reports, the causes and recommended preventive actions were considerably altered by the independent analysis and additional information from the facility. 130 reports have been entered in RIRAS; 9 misadministrations, 83 good catches, 3 anonymous good catches, and 35 earlier reported incidents from FY2005-14. 45% of the reported incidents occurred in the treatment delivery stages, 19% in on-treatment management, and 16% in pre-treatment verification. 80% of the good catches were found in the treatment delivery workflow. Majority of these incidents were due to inconsistent patient setup instructions or documentation, nonadherence to policies and procedures, lax time-out policy, distracted RTTs, and inadequate RTT staffing. Conclusion: RIRAS has identified many areas for improvement and elevated the quality and safety of radiation treatments in the VHA. We found that the ability to learn is significantly diminished when the analysts do not have the ability to request additional information.« less

  2. SU-E-T-255: Optimized Supine Craniospinal Irradiation with Image-Guided and Field Matched Beams

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jiang, Z; Holupka, E; Naughton, J

    2014-06-01

    Purpose: Conventional craniospinal irradiation (CSI) challenges include dose inhomogeneity at field junctions and position uncertainty due to the field divergence, particular for the two spinal fields. Here we outline a new supine CSI technique to address these difficulties. Methods: Patient was simulated in supine position. The cranial fields had isocenter at C2/C3 vertebral and were matched with 1st spinal field. Their inferior border was chosen to avoid the shoulder, as well as chin from the 1st spine field. Their collimator angles were dependent on asymmetry jaw setting of the 1st spinal field. With couch rotation, the spinal field gantry anglesmore » were adjusted to ensure, the inferior border of 1st and superior border of 2nd spinal fields were perpendicular to the table top. The radio-opaque wire position for the spinal junction was located initially by the light field from an anterior setup beam, and was finalized by the portal imaging of the 1st spinal field. With reference to the spinal junction wire, the fields were matched by positioning the isocenter of the 2nd spinal field. A formula was derived to optimize supine CSI treatment planning, by utilizing the relationship among the Yjaw setting, the spinal field gantry angles, cranial field collimator angles, and the spinal field isocenters location. The plan was delivered with portal imaging alignment for the both cranial and spinal junctions. Results: Utilizing this technique with matching beams, and conventional technique such as feathering and forwarding planning, a homogenous dose distribution was achieved throughout the entire CSI treatment volume including the spinal junction. Placing the spinal junction wire visualized in both spinal portals, allows for precise determination and verification of the appropriate match line of the spine fields. Conclusion: This technique of optimization supine CSI achieved a homogenous dose distributions and patient localization accuracy with image-guided and matched beams.« less

  3. SU-F-T-316: A Model to Deal with Dosimetric and Delivery Uncertainties in Radiotherapy Treatment Planning

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Haering, P; Lang, C; Splinter, M

    2016-06-15

    Purpose The conventional way of dealing with uncertainties resulting from dose calculation or beam delivery in IMRT, is to do verification measurements for the plan in question. Here we present an alternative based on recommendations given in the AAPM 142 report and treatment specific parameters that model the uncertainties for the plan delivery. Methods Basis of the model is the assignment of uncertainty parameters to all segment fields or control point sequences of a plan. The given field shape is analyzed for complexity, dose rate, number of MU, field size related output as well as factors for in/out field positionmore » and penumbra regions. Together with depth related uncertainties, a 3D matrix is generated by a projection algorithm. Patient anatomy is included as uncertainty CT data set as well. Therefore, object density is classified in 4 categories close to water, lung, bone and gradient regions with additional uncertainties. The result is then exported as a DICOM dose file by the software tool (written in IDL, Exelis), having the given resolution and target point. Results Uncertainty matrixes for several patient cases have been calculated and compared side by side in the planning system. The result is not quite always intuitive but it clearly indicates high and low uncertainties related to OARs and target volumes as well as to measured gamma distributions.ConclusionThe imported uncertainty datasets may help the treatment planner to understand the complexity of the treatment plan. He then might decide to change the plan to produce a more suited uncertainty distribution, e.g. by changing the beam angles the high uncertainty spots can be influenced or try to use another treatment setup, resulting in a plan with lower uncertainties. A next step could be to include such a model into the optimization algorithm to add a new dose uncertainty constraint.« less

  4. Three-dimensional analysis of the surface registration accuracy of electromagnetic navigation systems in live endoscopic sinus surgery.

    PubMed

    Chang, C M; Fang, K M; Huang, T W; Wang, C T; Cheng, P W

    2013-12-01

    Studies on the performance of surface registration with electromagnetic tracking systems are lacking in both live surgery and the laboratory setting. This study presents the efficiency in time of the system preparation as well as the navigational accuracy of surface registration using electromagnetic tracking systems. Forty patients with bilateral chronic paranasal pansinusitis underwent endoscopic sinus surgery after undergoing sinus computed tomography scans. The surgeries were performed under electromagnetic navigation guidance after the surface registration had been carried out on all of the patients. The intraoperative measurements indicate the time taken for equipment set-up, surface registration and surgical procedure, as well as the degree of navigation error along 3 axes. The time taken for equipment set-up, surface registration and the surgical procedure was 179 +- 23 seconds, 39 +- 4.8 seconds and 114 +- 36 minutes, respectively. A comparison of the navigation error along the 3 axes showed that the deviation in the medial-lateral direction was significantly less than that in the anterior-posterior and cranial-caudal directions. The procedures of equipment set-up and surface registration in electromagnetic navigation tracking are efficient, convenient and easy to manipulate. The system accuracy is within the acceptable ranges, especially on the medial-lateral axis.

  5. Inter- and Intrafraction Uncertainty in Prostate Bed Image-Guided Radiotherapy

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Huang, Kitty; Palma, David A.; Department of Oncology, University of Western Ontario, London

    2012-10-01

    Purpose: The goals of this study were to measure inter- and intrafraction setup error and prostate bed motion (PBM) in patients undergoing post-prostatectomy image-guided radiotherapy (IGRT) and to propose appropriate population-based three-dimensional clinical target volume to planning target volume (CTV-PTV) margins in both non-IGRT and IGRT scenarios. Methods and Materials: In this prospective study, 14 patients underwent adjuvant or salvage radiotherapy to the prostate bed under image guidance using linac-based kilovoltage cone-beam CT (kV-CBCT). Inter- and intrafraction uncertainty/motion was assessed by offline analysis of three consecutive daily kV-CBCT images of each patient: (1) after initial setup to skin marks, (2)more » after correction for positional error/immediately before radiation treatment, and (3) immediately after treatment. Results: The magnitude of interfraction PBM was 2.1 mm, and intrafraction PBM was 0.4 mm. The maximum inter- and intrafraction prostate bed motion was primarily in the anterior-posterior direction. Margins of at least 3-5 mm with IGRT and 4-7 mm without IGRT (aligning to skin marks) will ensure 95% of the prescribed dose to the clinical target volume in 90% of patients. Conclusions: PBM is a predominant source of intrafraction error compared with setup error and has implications for appropriate PTV margins. Based on inter- and estimated intrafraction motion of the prostate bed using pre- and post-kV-CBCT images, CBCT IGRT to correct for day-to-day variances can potentially reduce CTV-PTV margins by 1-2 mm. CTV-PTV margins for prostate bed treatment in the IGRT and non-IGRT scenarios are proposed; however, in cases with more uncertainty of target delineation and image guidance accuracy, larger margins are recommended.« less

  6. SU-E-J-103: Setup Errors Analysis by Cone-Beam CT (CBCT)-Based Imaged-Guided Intensity Modulated Radiotherapy for Esophageal Cancer

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Yang, H; Wang, W; Hu, W

    2014-06-01

    Purpose: To quantify setup errors by pretreatment kilovolt cone-beam computed tomography(KV-CBCT) scans for middle or distal esophageal carcinoma patients. Methods: Fifty-two consecutive middle or distal esophageal carcinoma patients who underwent IMRT were included this study. A planning CT scan using a big-bore CT simulator was performed in the treatment position and was used as the reference scan for image registration with CBCT. CBCT scans(On-Board Imaging v1. 5 system, Varian Medical Systems) were acquired daily during the first treatment week. A total of 260 CBCT scans was assessed with a registration clip box defined around the PTV-thorax in the reference scanmore » based on(nine CBCTs per patient) bony anatomy using Offline Review software v10.0(Varian Medical Systems). The anterior-posterior(AP), left-right(LR), superiorinferior( SI) corrections were recorded. The systematic and random errors were calculated. The CTV-to-PTV margins in each CBCT frequency was based on the Van Herk formula (2.5Σ+0.7σ). Results: The SD of systematic error (Σ) was 2.0mm, 2.3mm, 3.8mm in the AP, LR and SI directions, respectively. The average random error (σ) was 1.6mm, 2.4mm, 4.1mm in the AP, LR and SI directions, respectively. The CTV-to-PTV safety margin was 6.1mm, 7.5mm, 12.3mm in the AP, LR and SI directions based on van Herk formula. Conclusion: Our data recommend the use of 6 mm, 8mm, and 12 mm for esophageal carcinoma patient setup in AP, LR, SI directions, respectively.« less

  7. Poster — Thur Eve — 32: Stereotactic Body Radiation Therapy for Peripheral Lung Lesion: Treatment Planning and Quality Assurance

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wan, Shuying; Oliver, Michael; Wang, Xiaofang

    2014-08-15

    Stereotactic body radiation therapy (SBRT), due to its high precision for target localizing, has become widely used to treat tumours at various locations, including the lungs. Lung SBRT program was started at our institution a year ago. Eighteen patients with peripheral lesions up to 3 cm diameter have been treated with 48 Gy in 4 fractions. Based on four-dimensional computed tomography (4DCT) simulation, internal target volume (ITV) was delineated to encompass the respiratory motion of the lesion. A margin of 5 mm was then added to create the planning target volume (PTV) for setup uncertainties. There was no expansion frommore » gross tumour volume (GTV) to clinical target volume (CTV). Pinnacle 9.6 was used as the primary treatment planning system. Volumetric modulated arc therapy (VMAT) technique, with one or two coplanar arcs, generally worked well. For quality assurance (QA), each plan was exported to Eclipse 10 and dose calculation was repeated. Dose volume histograms (DVHs) of the targets and organs at risk (OARs) were then compared between the two treatment planning systems. Winston-Lutz tests were carried out as routine machine QA. Patient-specific QA included ArcCheck measurement with an insert, where an ionization chamber was placed at the centre to measure dose at the isocenter. For the first several patients, and subsequently for the plans with extremely strong modulation, Gafchromic film dosimetry was also employed. For each patient, a mock setup was scheduled prior to treatments. Daily pre- and post-CBCT were acquired for setup and assessment of intra-fractional motion, respectively.« less

  8. Dose delivered from Varian's CBCT to patients receiving IMRT for prostate cancer.

    PubMed

    Wen, Ning; Guan, Huaiqun; Hammoud, Rabih; Pradhan, Deepak; Nurushev, T; Li, Shidong; Movsas, Benjamin

    2007-04-21

    With the increased use of cone beam CT (CBCT) for daily patient setup, the accumulated dose from CBCT may be significantly higher than that from simulation CT or portal imaging. The objective of this work is to measure the dose from daily pelvic scans with fixed technical settings and collimations. CBCT scans were acquired in half-fan mode using a half bowtie and x-rays were delivered in pulsed-fluoro mode. The skin doses for seven prostate patients were measured on an IRB-approved protocol. TLD capsules were placed on the patient's skin at the central axis of three beams: AP, left lateral (Lt Lat) and right lateral (Rt Lat). To avoid the ring artefacts centred in the prostate, the treatment couch was dropped 3 cm from the patient's tattoo (central axis). The measured AP skin doses ranged 3-6 cGy for 20-33 cm separation. The larger the patient size the less the AP skin dose. Lateral doses did not change much with patient size. The Lt Lat dose was approximately 4.0 cGy, which was approximately 40% higher than the Rt Lat dose of approximately 2.6 cGy. To verify this dose asymmetry, surface doses on an IMRT QA phantom (oval shaped, 30 cm x 20 cm) were measured at the same three sites using TLD capsules with 3 cm table-drop. The dose asymmetry was due to: (1) kV source rotation which always starts from the patient's Lt Lat and ends at Lt Lat. Gantry rotation gets much slower near the end of rotation but dose rate stays constant and (2) 370 degrees scan rotation (10 degrees scan overlap on the Lt Lat side). In vivo doses were measured inside a Rando pelvic heterogeneous phantom using TLDs. The left hip (femoral head and neck) received the highest doses of approximately 10-11 cGy while the right hip received approximately 6-7 cGy. The surface and in vivo doses were also measured for phantoms at the central-axis setup. The difference was less than approximately 12% to the table-drop setup.

  9. Dose delivered from Varian's CBCT to patients receiving IMRT for prostate cancer

    NASA Astrophysics Data System (ADS)

    Wen, Ning; Guan, Huaiqun; Hammoud, Rabih; Pradhan, Deepak; Nurushev, T.; Li, Shidong; Movsas, Benjamin

    2007-04-01

    With the increased use of cone beam CT (CBCT) for daily patient setup, the accumulated dose from CBCT may be significantly higher than that from simulation CT or portal imaging. The objective of this work is to measure the dose from daily pelvic scans with fixed technical settings and collimations. CBCT scans were acquired in half-fan mode using a half bowtie and x-rays were delivered in pulsed-fluoro mode. The skin doses for seven prostate patients were measured on an IRB-approved protocol. TLD capsules were placed on the patient's skin at the central axis of three beams: AP, left lateral (Lt Lat) and right lateral (Rt Lat). To avoid the ring artefacts centred in the prostate, the treatment couch was dropped 3 cm from the patient's tattoo (central axis). The measured AP skin doses ranged 3-6 cGy for 20-33 cm separation. The larger the patient size the less the AP skin dose. Lateral doses did not change much with patient size. The Lt Lat dose was ~4.0 cGy, which was ~40% higher than the Rt Lat dose of ~2.6 cGy. To verify this dose asymmetry, surface doses on an IMRT QA phantom (oval shaped, 30 cm × 20 cm) were measured at the same three sites using TLD capsules with 3 cm table-drop. The dose asymmetry was due to: (1) kV source rotation which always starts from the patient's Lt Lat and ends at Lt Lat. Gantry rotation gets much slower near the end of rotation but dose rate stays constant and (2) 370° scan rotation (10° scan overlap on the Lt Lat side). In vivo doses were measured inside a Rando pelvic heterogeneous phantom using TLDs. The left hip (femoral head and neck) received the highest doses of ~10-11 cGy while the right hip received ~6-7 cGy. The surface and in vivo doses were also measured for phantoms at the central-axis setup. The difference was less than ~12% to the table-drop setup.

  10. Methodology to reduce 6D patient positional shifts into a 3D linear shift and its verification in frameless stereotactic radiotherapy

    NASA Astrophysics Data System (ADS)

    Sarkar, Biplab; Ray, Jyotirmoy; Ganesh, Tharmarnadar; Manikandan, Arjunan; Munshi, Anusheel; Rathinamuthu, Sasikumar; Kaur, Harpreet; Anbazhagan, Satheeshkumar; Giri, Upendra K.; Roy, Soumya; Jassal, Kanan; Kalyan Mohanti, Bidhu

    2018-04-01

    The aim of this article is to derive and verify a mathematical formulation for the reduction of the six-dimensional (6D) positional inaccuracies of patients (lateral, longitudinal, vertical, pitch, roll and yaw) to three-dimensional (3D) linear shifts. The formulation was mathematically and experimentally tested and verified for 169 stereotactic radiotherapy patients. The mathematical verification involves the comparison of any (one) of the calculated rotational coordinates with the corresponding value from the 6D shifts obtained by cone beam computed tomography (CBCT). The experimental verification involves three sets of measurements using an ArcCHECK phantom, when (i) the phantom was not moved (neutral position: 0MES), (ii) the position of the phantom shifted by 6D shifts obtained from CBCT (6DMES) from neutral position and (iii) the phantom shifted from its neutral position by 3D shifts reduced from 6D shifts (3DMES). Dose volume histogram and statistical comparisons were made between ≤ft< TPSCAL{\\text -}0MES \\right> and ≤ft< 3DMES{\\text -6DMES} \\right> . The mathematical verification was performed by a comparison of the calculated and measured yaw (γ°) rotation values, which gave a straight line, Y  =  1X with a goodness of fit as R 2  =  0.9982. The verification, based on measurements, gave a planning target volume receiving 100% of the dose (V100%) as 99.1  ±  1.9%, 96.3  ±  1.8%, 74.3  ±  1.9% and 72.6  ±  2.8% for the calculated treatment planning system values TPSCAL, 0MES, 3DMES and 6DMES, respectively. The statistical significance (p-values: paired sample t-test) of V100% were found to be 0.03 for the paired sample ≤ft< 3DMES{\\text -6DMES} \\right> and 0.01 for ≤ft< 0MES{\\text -TPSCAL} \\right> . In this paper, a mathematical method to reduce 6D shifts to 3D shifts is presented. The mathematical method is verified by using well-matched values between the measured and calculated γ°. Measurements done on the ArcCHECK phantom also proved that the proposed methodology is correct. The post-correction of the table position condition introduces a minimal spatial dose delivery error in the frameless stereotactic system, using a 6D motion enabled robotic couch. This formulation enables the reduction of 6D positional inaccuracies to 3D linear shifts, and hence allows the treatment of patients with frameless stereotactic radiosurgery by using only a 3D linear motion enabled couch.

  11. Methodology to reduce 6D patient positional shifts into a 3D linear shift and its verification in frameless stereotactic radiotherapy.

    PubMed

    Sarkar, Biplab; Ray, Jyotirmoy; Ganesh, Tharmarnadar; Manikandan, Arjunan; Munshi, Anusheel; Rathinamuthu, Sasikumar; Kaur, Harpreet; Anbazhagan, Satheeshkumar; Giri, Upendra K; Roy, Soumya; Jassal, Kanan; Mohanti, Bidhu Kalyan

    2018-03-22

    The aim of this article is to derive and verify a mathematical formulation for the reduction of the six-dimensional (6D) positional inaccuracies of patients (lateral, longitudinal, vertical, pitch, roll and yaw) to three-dimensional (3D) linear shifts. The formulation was mathematically and experimentally tested and verified for 169 stereotactic radiotherapy patients. The mathematical verification involves the comparison of any (one) of the calculated rotational coordinates with the corresponding value from the 6D shifts obtained by cone beam computed tomography (CBCT). The experimental verification involves three sets of measurements using an ArcCHECK phantom, when (i) the phantom was not moved (neutral position: 0MES), (ii) the position of the phantom shifted by 6D shifts obtained from CBCT (6DMES) from neutral position and (iii) the phantom shifted from its neutral position by 3D shifts reduced from 6D shifts (3DMES). Dose volume histogram and statistical comparisons were made between [Formula: see text] and [Formula: see text]. The mathematical verification was performed by a comparison of the calculated and measured yaw (γ°) rotation values, which gave a straight line, Y  =  1X with a goodness of fit as R 2   =  0.9982. The verification, based on measurements, gave a planning target volume receiving 100% of the dose (V100%) as 99.1  ±  1.9%, 96.3  ±  1.8%, 74.3  ±  1.9% and 72.6  ±  2.8% for the calculated treatment planning system values TPSCAL, 0MES, 3DMES and 6DMES, respectively. The statistical significance (p-values: paired sample t-test) of V100% were found to be 0.03 for the paired sample [Formula: see text] and 0.01 for [Formula: see text]. In this paper, a mathematical method to reduce 6D shifts to 3D shifts is presented. The mathematical method is verified by using well-matched values between the measured and calculated γ°. Measurements done on the ArcCHECK phantom also proved that the proposed methodology is correct. The post-correction of the table position condition introduces a minimal spatial dose delivery error in the frameless stereotactic system, using a 6D motion enabled robotic couch. This formulation enables the reduction of 6D positional inaccuracies to 3D linear shifts, and hence allows the treatment of patients with frameless stereotactic radiosurgery by using only a 3D linear motion enabled couch.

  12. Torque Loss After Miniscrew Placement: An In-Vitro Study Followed by a Clinical Trial.

    PubMed

    Migliorati, Marco; Drago, Sara; Barberis, Fabrizio; Schiavetti, Irene; Dalessandri, Domenico; Benedicenti, Stefano; Biavati, Armando Silvestrini

    2016-01-01

    To evaluate torque loss a week after insertion, both in an in vivo and an in vitro experimental setup were designed. In the in vivo setup a total of 29 miniscrews were placed in 20 patients who underwent orthodontic treatment. Maximum insertion torque (MIT) was evaluated at insertion time (T1). A week later, insertion torque was measured again by applying a quarter turn (T2); no load was applied on the screw during the first week. In the in vitro setup a total of 20 miniscrews were placed in pig rib bone samples. MIT was evaluated at insertion time (T1). Bone samples were kept in saline solution and controlled environment for a week during which the solution was refreshed every day. Afterwards, torque was measured again by applying a quarter turn (T2). The comparison of MIT over time was done calculating the percentage difference of the torque values between pre- and post-treatment and using the parametric two independent samples t-test or the non-parametric Mann-Whitney test. After a week unloaded miniscrews showed a mean loss of rotational torque of 36.3% and 40.9% in in vitro and in in vivo conditions, respectively. No statistical differences were found between the two different setups. Torque loss was observed after the first week in both study models; in vitro experimental setup provided a reliable study model for studying torque variation during the first week after insertion.

  13. Torque Loss After Miniscrew Placement: An In-Vitro Study Followed by a Clinical Trial

    PubMed Central

    Migliorati, Marco; Drago, Sara; Barberis, Fabrizio; Schiavetti, Irene; Dalessandri, Domenico; Benedicenti, Stefano; Biavati, Armando Silvestrini

    2016-01-01

    To evaluate torque loss a week after insertion, both in an in vivo and an in vitro experimental setup were designed. In the in vivo setup a total of 29 miniscrews were placed in 20 patients who underwent orthodontic treatment. Maximum insertion torque (MIT) was evaluated at insertion time (T1). A week later, insertion torque was measured again by applying a quarter turn (T2); no load was applied on the screw during the first week. In the in vitro setup a total of 20 miniscrews were placed in pig rib bone samples. MIT was evaluated at insertion time (T1). Bone samples were kept in saline solution and controlled environment for a week during which the solution was refreshed every day. Afterwards, torque was measured again by applying a quarter turn (T2). The comparison of MIT over time was done calculating the percentage difference of the torque values between pre- and post-treatment and using the parametric two independent samples t-test or the non-parametric Mann–Whitney test. After a week unloaded miniscrews showed a mean loss of rotational torque of 36.3% and 40.9% in in vitro and in in vivo conditions, respectively. No statistical differences were found between the two different setups. Torque loss was observed after the first week in both study models; in vitro experimental setup provided a reliable study model for studying torque variation during the first week after insertion. PMID:27386011

  14. Interfraction Displacement of Primary Tumor and Involved Lymph Nodes Relative to Anatomic Landmarks in Image Guided Radiation Therapy of Locally Advanced Lung Cancer

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Jan, Nuzhat; Balik, Salim; Hugo, Geoffrey D.

    Purpose: To analyze primary tumor (PT) and lymph node (LN) position changes relative to each other and relative to anatomic landmarks during conventionally fractionated radiation therapy for patients with locally advanced lung cancer. Methods and Materials: In 12 patients with locally advanced non-small cell lung cancer PT, LN, carina, and 1 thoracic vertebra were manually contoured on weekly 4-dimensional fan-beam CT scans. Systematic and random interfraction displacements of all contoured structures were identified in the 3 cardinal directions, and resulting setup margins were calculated. Time trends and the effect of volume changes on displacements were analyzed. Results: Three-dimensional displacement vectorsmore » and systematic/random interfraction displacements were smaller for carina than for vertebra both for PT and LN. For PT, mean (SD) 3-dimensional displacement vectors with carina-based alignment were 7 (4) mm versus 9 (5) mm with bony anatomy (P<.0001). For LN, smaller displacements were found with carina- (5 [3] mm, P<.0001) and vertebra-based (6 [3] mm, P=.002) alignment compared with using PT for setup (8 [5] mm). Primary tumor and LN displacements relative to bone and carina were independent (P>.05). Displacements between PT and bone (P=.04) and between PT and LN (P=.01) were significantly correlated with PT volume regression. Displacements between LN and carina were correlated with LN volume change (P=.03). Conclusions: Carina-based setup results in a more reproducible PT and LN alignment than bony anatomy setup. Considering the independence of PT and LN displacement and the impact of volume regression on displacements over time, repeated CT imaging even with PT-based alignment is recommended in locally advanced disease.« less

  15. SU-E-J-248: Comparative Study of Two Image Registration for Image-Guided Radiation Therapy in Esophageal Cancer

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Shang, K; Wang, J; Liu, D

    2014-06-01

    Purpose: Image-guided radiation therapy (IGRT) is one of the major treatment of esophageal cancer. Gray value registration and bone registration are two kinds of image registration, the purpose of this work is to compare which one is more suitable for esophageal cancer patients. Methods: Twenty three esophageal patients were treated by Elekta Synergy, CBCT images were acquired and automatically registered to planning kilovoltage CT scans according to gray value or bone registration. The setup errors were measured in the X, Y and Z axis, respectively. Two kinds of setup errors were analysed by matching T test statistical method. Results: Fourmore » hundred and five groups of CBCT images were available and the systematic and random setup errors (cm) in X, Y, Z directions were 0.35, 0.63, 0.29 and 0.31, 0.53, 0.21 with gray value registration, while 0.37, 0.64, 0.26 and 0.32, 0.55, 0.20 with bone registration, respectively. Compared with bone registration and gray value registration, the setup errors in X and Z axis have significant differences. In Y axis, both measurement comparison results of T value is 0.256 (P value > 0.05); In X axis, the T value is 5.287(P value < 0.05); In Z axis, the T value is −5.138 (P value < 0.05). Conclusion: Gray value registration is recommended in image-guided radiotherapy for esophageal cancer and the other thoracic tumors. Manual registration could be applied when it is necessary. Bone registration is more suitable for the head tumor and pelvic tumor department where composed of redundant interconnected and immobile bone tissue.« less

  16. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Klüter, Sebastian, E-mail: sebastian.klueter@med.uni-heidelberg.de; Schubert, Kai; Lissner, Steffen

    Purpose: The dosimetric verification of treatment plans in helical tomotherapy usually is carried out via verification measurements. In this study, a method for independent dose calculation of tomotherapy treatment plans is presented, that uses a conventional treatment planning system with a pencil kernel dose calculation algorithm for generation of verification dose distributions based on patient CT data. Methods: A pencil beam algorithm that directly uses measured beam data was configured for dose calculation for a tomotherapy machine. Tomotherapy treatment plans were converted into a format readable by an in-house treatment planning system by assigning each projection to one static treatmentmore » field and shifting the calculation isocenter for each field in order to account for the couch movement. The modulation of the fluence for each projection is read out of the delivery sinogram, and with the kernel-based dose calculation, this information can directly be used for dose calculation without the need for decomposition of the sinogram. The sinogram values are only corrected for leaf output and leaf latency. Using the converted treatment plans, dose was recalculated with the independent treatment planning system. Multiple treatment plans ranging from simple static fields to real patient treatment plans were calculated using the new approach and either compared to actual measurements or the 3D dose distribution calculated by the tomotherapy treatment planning system. In addition, dose–volume histograms were calculated for the patient plans. Results: Except for minor deviations at the maximum field size, the pencil beam dose calculation for static beams agreed with measurements in a water tank within 2%/2 mm. A mean deviation to point dose measurements in the cheese phantom of 0.89% ± 0.81% was found for unmodulated helical plans. A mean voxel-based deviation of −0.67% ± 1.11% for all voxels in the respective high dose region (dose values >80%), and a mean local voxel-based deviation of −2.41% ± 0.75% for all voxels with dose values >20% were found for 11 modulated plans in the cheese phantom. Averaged over nine patient plans, the deviations amounted to −0.14% ± 1.97% (voxels >80%) and −0.95% ± 2.27% (>20%, local deviations). For a lung case, mean voxel-based deviations of more than 4% were found, while for all other patient plans, all mean voxel-based deviations were within ±2.4%. Conclusions: The presented method is suitable for independent dose calculation for helical tomotherapy within the known limitations of the pencil beam algorithm. It can serve as verification of the primary dose calculation and thereby reduce the need for time-consuming measurements. By using the patient anatomy and generating full 3D dose data, and combined with measurements of additional machine parameters, it can substantially contribute to overall patient safety.« less

  17. A novel and fast method for proton range verification using a step wedge and 2D scintillator.

    PubMed

    Shen, Jiajian; Allred, Bryce C; Robertson, Daniel G; Liu, Wei; Sio, Terence T; Remmes, Nicholas B; Keole, Sameer R; Bues, Martin

    2017-09-01

    To implement and evaluate a novel and fast method for proton range verification by using a planar scintillator and step wedge. A homogenous proton pencil beam plan with 35 energies was designed and delivered to a 2D flat scintillator with a step wedge. The measurement was repeated 15 times (3 different days, 5 times per day). The scintillator image was smoothed, the Bragg peak and distal fall off regions were fitted by an analytical equation, and the proton range was calculated using simple trigonometry. The accuracy of this method was verified by comparing the measured ranges to those obtained using an ionization chamber and a scanning water tank, the gold standard. The reproducibility was evaluated by comparing the ranges over 15 repeated measurements. The sensitivity was evaluated by delivering to same beam to the system with a film inserted under the wedge. The range accuracy of all 35 proton energies measured over 3 days was within 0.2 mm. The reproducibility in 15 repeated measurements for all 35 proton ranges was ±0.045 mm. The sensitivity to range variation is 0.1 mm for the worst case. This efficient procedure permits measurement of 35 proton ranges in less than 3 min. The automated data processing produces results immediately. The setup of this system took less than 5 min. The time saving by this new method is about two orders of magnitude when compared with the time for water tank range measurements. A novel method using a scintillator with a step wedge to measure the proton range was implemented and evaluated. This novel method is fast and sensitive, and the proton range measured by this method was accurate and highly reproducible. © 2017 American Association of Physicists in Medicine.

  18. AXAF-I Low Intensity-Low Temperature (LILT) Testing of the Development Verification Test (DVT) Solar Panel

    NASA Technical Reports Server (NTRS)

    Alexander, Doug; Edge, Ted; Willowby, Doug

    1998-01-01

    The planned orbit of the AXAF-I spacecraft will subject the spacecraft to both short, less than 30 minutes for solar and less than 2 hours for lunar, and long earth eclipses and lunar eclipses with combined conjunctive duration of up to 3 to 4 hours. Lack of proper Electrical Power System (EPS) conditioning prior to eclipse may cause loss of mission. To avoid this problem, for short eclipses, it is necessary to off-point the solar array prior to or at the beginning of the eclipse to reduce the battery state of charge (SOC). This yields less overcharge during the high charge currents at sun entry. For long lunar eclipses, solar array pointing and load scheduling must be tailored for the profile of the eclipse. The battery SOC, loads, and solar array current-voltage (I-V) must be known or predictable to maintain the bus voltage within acceptable range. To address engineering concerns about the electrical performance of the AXAF-I solar array under Low Intensity and Low Temperature (LILT) conditions, Marshall Space Flight Center (MSFC) engineers undertook special testing of the AXAF-I Development Verification Test (DVT) solar panel in September-November 1997. In the test the DVT test panel was installed in a thermal vacuum chamber with a large view window with a mechanical "flapper door". The DVT test panel was "flash" tested with a Large Area Pulse Solar Simulator (LAPSS) at various fractional sun intensities and panel (solar cell) temperatures. The testing was unique with regards to the large size of the test article and type of testing performed. The test setup, results, and lessons learned from the testing will be presented.

  19. Clinical implementation and failure mode and effects analysis of HDR skin brachytherapy using Valencia and Leipzig surface applicators.

    PubMed

    Sayler, Elaine; Eldredge-Hindy, Harriet; Dinome, Jessie; Lockamy, Virginia; Harrison, Amy S

    2015-01-01

    The planning procedure for Valencia and Leipzig surface applicators (VLSAs) (Nucletron, Veenendaal, The Netherlands) differs substantially from CT-based planning; the unfamiliarity could lead to significant errors. This study applies failure modes and effects analysis (FMEA) to high-dose-rate (HDR) skin brachytherapy using VLSAs to ensure safety and quality. A multidisciplinary team created a protocol for HDR VLSA skin treatments and applied FMEA. Failure modes were identified and scored by severity, occurrence, and detectability. The clinical procedure was then revised to address high-scoring process nodes. Several key components were added to the protocol to minimize risk probability numbers. (1) Diagnosis, prescription, applicator selection, and setup are reviewed at weekly quality assurance rounds. Peer review reduces the likelihood of an inappropriate treatment regime. (2) A template for HDR skin treatments was established in the clinic's electronic medical record system to standardize treatment instructions. This reduces the chances of miscommunication between the physician and planner as well as increases the detectability of an error. (3) A screen check was implemented during the second check to increase detectability of an error. (4) To reduce error probability, the treatment plan worksheet was designed to display plan parameters in a format visually similar to the treatment console display, facilitating data entry and verification. (5) VLSAs are color coded and labeled to match the electronic medical record prescriptions, simplifying in-room selection and verification. Multidisciplinary planning and FMEA increased detectability and reduced error probability during VLSA HDR brachytherapy. This clinical model may be useful to institutions implementing similar procedures. Copyright © 2015 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  20. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Conroy, L; Quirk, S; Smith, WL

    Purpose: We used Real-Time Position Management (RPM) to evaluate breath hold amplitude and variability when gating with a visually monitored deep inspiration breath hold technique (VM-DIBH) with retrospective cine image chest wall position verification. Methods: Ten patients with left-sided breast cancer were treated using VM-DIBH. Respiratory motion was passively collected once weekly using RPM with the marker block positioned at the xiphoid process. Cine images on the tangent medial field were acquired on fractions with RPM monitoring for retrospective verification of chest wall position during breath hold. The amplitude and duration of all breath holds on which treatment beams weremore » delivered were extracted from the RPM traces. Breath hold position coverage was evaluated for symmetric RPM gating windows from ± 1 to 5 mm centered on the average breath hold amplitude of the first measured fraction as a baseline. Results: The average (range) breath hold amplitude and duration was 18 mm (3–36 mm) and 19 s (7–34 s). The average (range) of amplitude standard deviation per patient over all breath holds was 2.7 mm (1.2–5.7 mm). With the largest allowable RPM gating window (± 5 mm), 4 of 10 VM-DIBH patients would have had ≥ 10% of their breath hold positions excluded by RPM. Cine verification of the chest wall position during the medial tangent field showed that the chest wall was greater than 5 mm from the baseline in only 1 out of 4 excluded patients. Cine images verify the chest wall/breast position only, whether this variation is acceptable in terms of heart sparing is a subject of future investigation. Conclusion: VM-DIBH allows for greater breath hold amplitude variability than using a 5 mm gating window with RPM, while maintaining chest wall positioning accuracy within 5 mm for the majority of patients.« less

Top