Science.gov

Sample records for photon multileaf collimator

  1. Validation of modulated electron radiotherapy delivered with photon multileaf collimation

    NASA Astrophysics Data System (ADS)

    Klein, Eric E.

    There is a challenge in radiotherapy to treat shallow targets due to the inability to provide dose heterogeneity while simultaneously minimizing dose to distal critical organs. There is a niche for Modulated Electron Radiotherapy (MERT) to complement a photon IMRT program. Disease sites such as post-mastectomy chest wall, and subcutaneous lymphoma of the scalp, etc. are better suited for modulated electrons rather than photons, or perhaps a combination. Inherent collimation systems are not conducive for electron beam delivery (in lieu of extended applicators), nor do commercial treatment planning systems model electrons collimated without applicators. The purpose of this study is to evaluate modulation of electrons by inherent photon multileaf collimators, and calculated and optimized by means of Monte Carlo. Modulated electron radiotherapy (MERT) evaluation was conducted with a Trilogy 120 leaf MLC for 6-20 MeV. To provide a sharp penumbra, modulated beams were delivered with short SSDs (70-85cm). Segment widths (SW) ranging from 1 to 10cm were configured for delivery and planning, using BEAMnrc MC code with 109 particles, and DOSXYZnrc calculations. Calculations were set with: voxel size 0.2 x 0.2 x 0.1cm3, and photon/electron transport energy cutoffs of 0.01 MeV/0.521 MeV. Dosimetry was performed with film and micro chambers. Calculated and measured data were analyzed in MatLab. Once validation of static fields was successfully completed, modulated portals (segmented and dynamic) were configured for treatment and calculations. Optimization for target coverage and OAR sparing was achieved by choosing energies according to target depth, and SW according to spatial coverage. Intensity for each segment was optimized by MC methods. Beam sharpness (penumbra) degraded with: decreasing energy and SW, and increasing SSD. PDD decreased significantly with decreasing SW. We have demonstrated excellent calculation/measurement agreement (<3mm). Equal dose profiles were

  2. Matching of electron and photon beams with a multi-leaf collimator.

    PubMed

    Karlsson, M; Zackrisson, B

    1993-12-01

    Multi-leaf collimators (MLCs) are offered as an accessory to many accelerators for radiation therapy. However, beam edges generated with these collimators are not as smooth as can be achieved with individually made blocks. The clinical drawbacks and benefits of this ripple were evaluated both for single field treatments and for combined adjacent fields of different beam qualities. In this investigation the MLC-collimated beams of the MM50 racetrack microtron were studied. The distance between the field edge and the 90% isodose was measured at the reference depth for four beam qualities (20 MV photons and 10, 20 and 50 MeV electrons). This distance was found to vary from approximately 6 mm for straight beam edges (i.e., all collimator leaves aligned) to approximately 2 mm from the tip of the leaves for a saw-tooth shaped beam edge. The over- and under-dosage in the joint between combined adjacent fields was found to be typically +/- 10% in small volumes. Improved clinical techniques using adjacent photon and electron fields with the same isocentre and source position (without moving the gantry) have been developed. For treatments of the breast, including the mammary chain, a uniform dose distribution was created with special attention given to the irradiation of the heart and lung outside the target volume. A method for head and neck treatments was optimised to give uniform dose distribution in the joint between the photon and electron fields and a method of treating the mediastinum, including the chest wall in front of the left lung, was analysed with respect to dose uniformity in the tumour and shielding of the lung.

  3. Delivery of modulated electron beams with conventional photon multi-leaf collimators

    NASA Astrophysics Data System (ADS)

    Klein, Eric E.; Mamalui-Hunter, Maria; Low, Daniel A.

    2009-01-01

    Electron beam radiotherapy is an accepted method to treat shallow tumors. However, modulation of electrons to customize dose distributions has not readily been achieved. Studies of bolus and tertiary collimation systems have been met with limitations. We pursue the use of photon multi-leaf collimators (MLC) for modulated electron radiotherapy (MERT) to achieve customized distributions for potential clinical use. As commercial planning systems do not support the use of MLC with electrons, planning was conducted using Monte Carlo calculations. Segmented and dynamic modulated delivery of multiple electron segments was configured, calculated and delivered for validation. Delivery of electrons with segmented or dynamic leaf motion was conducted. A phantom possessing an idealized stepped target was planned and optimized with subsequent validation by measurements. Finally, clinical treatment plans were conducted for post-mastectomy and cutaneous lymphoma of the scalp using forward optimization techniques. Comparison of calculations and measurements was successful with agreement of ±2%/2 mm for the energies, segment sizes, depths tested for delivered segments for the dynamic and segmented delivery. Clinical treatment plans performed provided optimal dose coverage of the target while sparing distal organs at risk. Execution of plans using an anthropomorphic phantom to ensure safe and efficient delivery was conducted. Our study validates that MERT is not only possible using the photon MLC, but the efficient and safe delivery inherent with the dynamic delivery provides an ideal technique for shallow tumor treatment.

  4. Design of a fast multileaf collimator for radiobiological optimized IMRT with scanned beams of photons, electrons, and light ions

    SciTech Connect

    Svensson, Roger; Larsson, Susanne; Gudowska, Irena; Holmberg, Rickard; Brahme, Anders

    2007-03-15

    Intensity modulated radiation therapy is rapidly becoming the treatment of choice for most tumors with respect to minimizing damage to the normal tissues and maximizing tumor control. Today, intensity modulated beams are most commonly delivered using segmental multileaf collimation, although an increasing number of radiation therapy departments are employing dynamic multileaf collimation. The irradiation time using dynamic multileaf collimation depends strongly on the nature of the desired dose distribution, and it is difficult to reduce this time to less than the sum of the irradiation times for all individual peak heights using dynamic leaf collimation [Svensson et al., Phys. Med. Biol. 39, 37-61 (1994)]. Therefore, the intensity modulation will considerably increase the total treatment time. A more cost-effective procedure for rapid intensity modulation is using narrow scanned photon, electron, and light ion beams in combination with fast multileaf collimator penumbra trimming. With this approach, the irradiation time is largely independent of the complexity of the desired intensity distribution and, in the case of photon beams, may even be shorter than with uniform beams. The intensity modulation is achieved primarily by scanning of a narrow elementary photon pencil beam generated by directing a narrow well focused high energy electron beam onto a thin bremsstrahlung target. In the present study, the design of a fast low-weight multileaf collimator that is capable of further sharpening the penumbra at the edge of the elementary scanned beam has been simulated, in order to minimize the dose or radiation response of healthy tissues. In the case of photon beams, such a multileaf collimator can be placed relatively close to the bremsstrahlung target to minimize its size. It can also be flat and thin, i.e., only 15-25 mm thick in the direction of the beam with edges made of tungsten or preferably osmium to optimize the sharpening of the penumbra. The low height of

  5. MCNPX simulation of a multileaf collimator

    SciTech Connect

    Poenisch, Falk; Titt, Uwe; Kry, Stephen F.; Vassiliev, Oleg N.; Mohan, Radhe

    2006-02-15

    Our purpose in this work was to validate a very detailed Monte Carlo model, developed in MCNPX, of a Millennium 120 multileaf collimator integrated into a Varian Clinac 21EX treatment head. The Monte Carlo results were compared with measurements for both the 6-MV and 18-MV photon modes. The following comparisons were performed: depth-dose curves, lateral profiles, multileaf collimator leakage, the tongue-and-grove test, and the round leaf-end test. The good agreement between the Monte Carlo simulations and measurements showed that our model is accurate. Consequently, the benchmarks provided by our study can be used in future Monte Carlo studies.

  6. Characterization of megavoltage electron beams delivered through a photon multi-leaf collimator (pMLC)

    NASA Astrophysics Data System (ADS)

    du Plessis, F. C. P.; Leal, A.; Stathakis, S.; Xiong, W.; Ma, C.-M.

    2006-04-01

    A study is presented that characterizes megavoltage electron beams delivered through an existing double-focused photon multi-leaf collimator (pMLC) using film measurements in a solid water phantom. Machine output stability and linearity were evaluated as well as the effect of source-to-surface distance (SSD) and field size on the penumbra for electron energies between 6 and 18 MeV over an SSD range of 60-100 cm. Penumbra variations as a function of field size, depth of measurement and the influence of the jaws were also studied. Field abutment, field flatness and target coverage for segmented beams were also addressed. The measured field size for electrons transported through the pMLC was the same as that for an x-ray beam up to SSDs of 70 cm. At larger SSD, the lower energy electron fields deviated from the projected field. Penumbra data indicated that 60 cm SSD was the most favourable treatment distance. Backprojection of P20-80 penumbra data yielded a virtual source position located at 98.9 cm from the surface for 18 MeV electrons. For 6 MeV electrons, the virtual source position was at a distance of 82.6 cm. Penumbra values were smaller for small beam slits and reached a near-constant value for field widths larger than 5 cm. The influence of the jaws had a small effect on the penumbra. The R90 values ranged from 1.4 to 4.8 cm between 6 and 21 MeV as measured at 60 cm SSD for a 9 × 9 cm2 field. Uniformity and penumbra improvement could be demonstrated using weighted abutted fields especially useful for small segments. No detectable electron leakage through the pMLC was observed. Bremsstrahlung measurements taken at 60 cm SSD for a 9 × 9 cm2 field as shaped by the pMLC compared within 1% to bremsstrahlung measurements taken at 100 cm SSD for a 10 × 10 cm2 electron applicator field at 100 cm SSD.

  7. Peripheral dose measurements for 6 and 18 MV photon beams on a linear accelerator with multileaf collimator

    SciTech Connect

    Mazonakis, Michalis; Zacharopoulou, Fotini; Varveris, Haralambos; Damilakis, John

    2008-10-15

    Peripheral dose (PD) to critical structures outside treatment volume is of clinical importance. The aim of the current study was to estimate PD on a linear accelerator equipped with multileaf collimator (MLC). Dose measurements were carried out using an ionization chamber embedded in a water phantom for 6 and 18 MV photon beams. PD values were acquired for field sizes from 5x5 to 20x20 cm{sup 2} in increments of 5 cm at distances up to 24 cm from the field edge. Dose data were obtained at two collimator orientations where the measurement points are shielded by MLC and jaws. The variation of PD with the source to skin distance (SSD), depth, and lateral displacement of the measurement point was evaluated. To examine the dependence of PD upon the tissue thickness at the entrance point of the beam, scattered dose was measured using thermoluminescent dosemeters placed on three anthropomorphic phantoms simulating 5- and 10-year-old children and an average adult patient. PD from 6 MV photons varied from 0.13% to 6.75% of the central-axis maximum dose depending upon the collimator orientation, extent of irradiated area, and distance from the treatment field. The corresponding dose range from 18 MV x rays was 0.09% to 5.61%. The variation of PD with depth and with lateral displacements up to 80% of the field dimension was very small. The scattered dose from both photon beams increased with the increase of SSD or tissue thickness along beam axis. The presented dosimetric data set allows the estimation of scattered dose outside the primary beam.

  8. Modeling of a multileaf collimator

    NASA Astrophysics Data System (ADS)

    Kim, Siyong

    A comprehensive physics model of a multileaf collimator (MLC) field for treatment planning was developed. Specifically, an MLC user interface module that includes a geometric optimization tool and a general method of in- air output factor calculation were developed. An automatic tool for optimization of MLC conformation is needed to realize the potential benefits of MLC. It is also necessary that a radiation therapy treatment planning (RTTP) system is capable of modeling MLC completely. An MLC geometric optimization and user interface module was developed. The planning time has been reduced significantly by incorporating the MLC module into the main RTTP system, Radiation Oncology Computer System (ROCS). The dosimetric parameter that has the most profound effect on the accuracy of the dose delivered with an MLC is the change in the in-air output factor that occurs with field shaping. It has been reported that the conventional method of calculating an in-air output factor cannot be used for MLC shaped fields accurately. Therefore, it is necessary to develop algorithms that allow accurate calculation of the in-air output factor. A generalized solution for an in-air output factor calculation was developed. Three major contributors of scatter to the in-air output-flattening filter, wedge, and tertiary collimator-were considered separately. By virtue of a field mapping method, in which a source plane field determined by detector's eye view is mapped into a detector plane field, no additional dosimetric data acquisition other than the standard data set for a range of square fields is required for the calculation of head scatter. Comparisons of in-air output factors between calculated and measured values show a good agreement for both open and wedge fields. For rectangular fields, a simple equivalent square formula was derived based on the configuration of a linear accelerator treatment head. This method predicts in-air output to within 1% accuracy. A two

  9. Dosimetric evaluation of 120-leaf multileaf collimator in a Varian linear accelerator with 6-MV and 18-MV photon beams

    PubMed Central

    Mohan, R.; Jayesh, K.; Joshi, R. C.; Al-idrisi, Maha; Narayanamurthy, P.; Majumdar, Saroj Kumar Das

    2008-01-01

    In this study the dosimetric characteristics of 120-leaf multileaf collimators (MLCs) were evaluated for 6-MV and 18-MV photon beams. The dose rate, percentage depth dose, surface dose, dose in the build-up region, beam profile, flatness, symmetry, and penumbra width were measured using three field-defining methods: (i) ‘Jaw only’, (ii) ‘MLC only’, and (iii) ‘MLC+Jaw’. Analysis of dose rate shows that the dose rate for ‘MLC only’ field was higher than that for ‘Jaw only” and ‘MLC+Jaw’ fields in both the energies. The ‘percentage of difference’ of dose rates between ‘MLC only’ and ‘MLC+Jaw’ was (0.9% to 4.4%) and (1.14% to 7%) for 6 MV and 18 MV respectively. The surface dose and dose in the build-up region were more pronounced for ‘MLC only’ fields for both energies, and no significant difference was found in percentage depth dose beyond dmax for both energies. Beam profiles show that flatness and symmetry for both the energies were less than the 3%. The penumbra width for ‘MLC only’ field was more than that for the other two field-defining methods by (1 to 2 mm) and (0.8 to 1.3 mm) for 6-MV and 18-MV photon beams respectively. Analysis of ‘width of 50% dose level’ of the beam profiles at dmax to reflect the field size shows 1 to 2 mm more for 6-MV photons and 2.2 to 2.4 mm morefor 18-MV photons for ‘MLC only’ fields. The results of this study suggest that the characteristics of 120-leaf MLC system with 6 MV and 18 MV are same in all aspects except the surface dose, penumbra, dose in the build-up region, and width of 50% dose levels. PMID:19893701

  10. Matching extended-SSD electron beams to multileaf collimated photon beams in the treatment of head and neck cancer

    SciTech Connect

    Steel, Jared; Stewart, Allan; Satory, Philip

    2009-09-15

    Purpose: Matching the penumbra of a 6 MeV electron beam to the penumbra of a 6 MV photon beam is a dose optimization challenge, especially when the electron beam is applied from an extended source-to-surface distance (SSD), as in the case of some head and neck treatments. Traditionally low melting point alloy blocks have been used to define the photon beam shielding over the spinal cord region. However, these are inherently time consuming to construct and employ in the clinical situation. Multileaf collimators (MLCs) provide a fast and reproducible shielding option but generate geometrically nonconformal approximations to the desired beam edge definition. The effects of substituting Cerrobend for the MLC shielding mode in the context of beam matching with extended-SSD electron beams are the subject of this investigation. Methods: Relative dose beam data from a Varian EX 2100 linear accelerator were acquired in a water tank under the 6 MeV electron beam at both standard and extended-SSD and under the 6 MV photon beam defined by Cerrobend and a number of MLC stepping regimes. The effect of increasing the electron beam SSD on the beam penumbra was assessed. MLC stepping was also assessed in terms of the effects on both the mean photon beam penumbra and the intraleaf dose-profile nonuniformity relative to the MLC midleaf. Computational techniques were used to combine the beam data so as to simulate composite relative dosimetry in the water tank, allowing fine control of beam abutment gap variation. Idealized volumetric dosimetry was generated based on the percentage depth-dose data for the beam modes and the abutment geometries involved. Comparison was made between each composite dosimetry dataset and the relevant ideal dosimetry dataset by way of subtraction. Results: Weighted dose-difference volume histograms (DDVHs) were produced, and these, in turn, summed to provide an overall dosimetry score for each abutment and shielding type/angle combination. Increasing the

  11. Multileaf collimator for Coline medical accelerators

    NASA Astrophysics Data System (ADS)

    Harasimowicz, Janusz; Plebański, Grzegorz; Sajna, Krzysztof

    2008-01-01

    Multileaf collimator (MLC) allows advanced field shaping for radiation therapy delivered with medical accelerators. In this paper theoretical considerations and scientific studies of a new MLC design are described. Considered multileaf collimator model comprises of a multiplicity of tungsten leaves of 1 cm width projected at isocenter plane. To ensure compatibility of a new MLC solution with different accelerator types as well as to assure high reliability in irradiated environment and presence of strong magnetic field, a complex and independent control system had to be developed. It comprises of two modules - one placed in the accelerator treatment head and the other one placed in the control room. Both of them ensure high reliability and treatment quality while working in harsh conditions. Mechanical design and leaf shape optimization algorithm based on a ray tracing method are also described in details. Adapted solutions allowed providing minimized and uniform radiation penumbrae in the full range of leaves positions which is crucial for modern advanced radiotherapy.

  12. A multileaf collimator for neutron radiation therapy

    NASA Astrophysics Data System (ADS)

    Farr, J. B.; Maughan, R. L.; Yudelev, M.; Forman, J. D.; Blosser, E. J.; Horste, T.

    2001-12-01

    A multi-leaf collimator (MLC) has been designed for installation on the super-conducting cyclotron at the Gershenson Radiation Oncology Center. This MLC will replace the existing multi-rod collimator and the increased efficiency thus achieved should allow for a 50% increase in the number of patients treated. A study of the penumbra region of the neutron beam with focused and unfocused collimator leaves has been completed, together with activation measurements in steel and tungsten. Results of these studies were used to finalize the collimator leaf design. A steel collimator leaf with a 5 mm projection at the isocenter and a wedge shaped section has been chosen, to provide beam divergence in the direction perpendicular to the leaf motion. The leaf profile is "stepped" to prevent neutron leakage. The rationale for this leaf design is discussed. The overall design of the collimator system and the incorporation of a remote wedge-changing device will be presented. Each leaf is positioned using a stepping motor; the leaf position is independently confirmed using an optical system incorporating a coherent fiber optic and a CCD camera. The control system is being designed to allow for the implementation of intensity modulated neutron radiation therapy (IMNRT).

  13. Application of multileaf collimator in breast cancer radiation techniques

    NASA Astrophysics Data System (ADS)

    Janiszewska, Marzena; Dupla, Dorota; Nowakowski, Grzegorz

    2004-07-01

    Modern radiation therapy tools allow a precise delivery of a high dose to a target area (so-called planning target volume -- PTV) and spare, at the same time, critical organs in the vicinity of cancerous lesions. One of the tools of conformal therapy is a multi-leaf collimator, which provides the opportunity to optimally adjust the therapeutic field to the tumor area. More difficult areas for radiation therapy include: mamma, after BCT, and chest after mammectomy with regional lymph nodes. The objective of the study is to present technical and physical aspects of breast carcinoma irradiation when applying a multi-leaf collimator. The following techniques were applied: (a) the isocentric technique of tangent fields (from two to four) for the mamma after BCT; (b) the method of a common isocenter, for the areas of mamma and for regional lymph nodes; (c) the technique of complementary photon + electron fields, for the area of chest after mammectomy and lymph nodes. The presented techniques were implemented as standard procedures in the preparation of breast carcinoma radiation treatment in the Lower Silesian Oncology Center.

  14. Modeling multileaf collimators with the PEREGRINE Monte Carlo

    SciTech Connect

    Albright, N; Fujino, D H; J Wieczorek

    1999-03-01

    Multileaf collimators (MLCs) are becoming increasingly important for beam shaping and intensity modulated radiation therapy (IMRT). Their unique design can introduce subtle effects in the patient/phantom dose distribution. The PEREGRINE 3D Monte Carlo dose calculation system predicts dose by implementing a full Monte Carlo simulation of the beam delivery and patient/phantom system. As such, it provides a powerful tool to explore dosimetric effects of MLC designs. We have installed a new MLC modeling package into PEREGRINE. This package simulates full photon and electron transport in the MLC and includes tongue-and-groove construction and curved or straight leaf ends in the leaf shape geometry. We tested the accuracy of the PEREGRINE MLC package by comparing PEREGRINE predictions with ion chamber, diode, and photographic film measurements taken with a Varian 2 1 OOC using 6 and 18 MV photon beams. Profile and depth dose measurements were made for the MLC configured into annulus and comb patterns. In all cases, PEREGRINE modeled these measurements to within experimental uncertainties. Our results demonstrate PEREGRINE's accuracy for modeling MLC characteristics, and suggest that PEREGRINE would be an ideal tool to explore issues such as (1) underdosing between leaves due to the ''tongue-and-groove'' effect when dose from multiple MLC patterns are added together, (2) radiation leakage in the bullnose region, and (3) dose under a single leaf due to scatter in the patient.

  15. Siemens Multileaf Collimator Characterization and Quality Assurance Approaches for Intensity-Modulated Radiotherapy

    SciTech Connect

    Bayouth, John E.

    2008-05-01

    Application of the multileaf collimator (MLC) has evolved from replacing blocks to create treatment fields to creating photon fluence modulation for intensity-modulated radiotherapy (IMRT). Multileaf collimator system performance requirements are far more stringent for such applications and will require increased performance for future applications, such as motion tracking. This article reviews Siemens MLC systems, including a technical description and dosimetric characteristics of 56-, 82-, and 160-leaf designs. Routine quality assurance of MLC for IMRT necessitates frequent and critical assessment of MLC leaf position calibration errors that can present in many different ways (e.g., accuracy, reproducibility, longevity, hysteresis, and collimator/gantry angle dependencies). Several techniques for measuring these errors are presented, along with qualitative and quantitative techniques for analyzing results. In particular, increased accuracy of leaf position measurement at variable gantry angles is enabled by spatial transformations to electronic portal imaging device position quantified by calibration protocols introduced with megavoltage cone beam. Measured values of X-ray transmission (intra-leaf, inter-leaf, and through abutting leaf pairs) and penumbra (leaf end, leaf tongue, leaf groove) are presented with an evaluation of their characterization by a treatment-planning system. The dosimetric impact of planning system model inadequacies is demonstrated for collimator scatter, dose profile values within 30 mm of the field edge, and the resultant effect demonstrated on clinical cases. Finally, a description of automated quality assurance delivery, analysis, and calibration protocols applicable for the specific vendor's system is provided.

  16. A dosimetric comparison of two multileaf collimator designs.

    PubMed

    Killoran, J H; Giraud, J Y; Chin, L

    2002-08-01

    We present the results of measurements designed to compare two different multileaf collimator (MLC) designs using a novel evaluation technique. The MLC designs evaluated were: a "single-focused" MLC (SF-MLC) mounted below the jaws, and a "double-focused" MLC, which is a complete replacement for the lower jaws. The ability of each MLC to conform isodose lines to a prescribed field edge (PFE) was evaluated using film dosimetry. Circular fields, centered on axis and off axis, were used because they produce a range of "angles of approach" between the MLC leaves and the PFE. They also have the advantage that for an ideal field shaping system the resulting isodoses are concentric perfect circles, a well-defined basis for evaluation. The amplitude of the oscillations of the 50% isodose line about the PFE and the penumbra width as determined by the 20%, 80%, and 90% isodose lines was examined. We observe that the 50% isodose line oscillates around the PFE with greater amplitude for SF-MLC. We attribute this, at least in part, to the rounded ends of the SF-MLC leaves. However, the SF-MLC has a noticeably sharper penumbra, which we attribute to its position further from the source. We conclude that these results are relevant for accurate dosimetric modeling of these devices. PMID:12201422

  17. High-resolution field shaping utilizing a masked multileaf collimator.

    PubMed

    Williams, P C; Cooper, P

    2000-08-01

    Multileaf collimators (MLCs) have become an important tool in the modern radiotherapy department. However, the current limit of resolution (1 cm at isocentre) can be too coarse for acceptable shielding of all fields. A number of mini- and micro-MLCs have been developed, with thinner leaves to achieve approved resolution. Currently however, such devices are limited to modest field sizes and stereotactic applications. This paper proposes a new method of high-resolution beam collimation by use of a tertiary grid collimator situated below the conventional MLC. The width of each slit in the grid is a submultiple of the MLC width. A composite shaped field is thus built up from a series of subfields, with the main MLC defining the length of each strip within each subfield. Presented here are initial findings using a prototype device. The beam uniformity achievable with such a device was examined by measuring transmission profiles through the grid using a diode. Profiles thus measured were then copied and superposed to generate composite beams, from which the uniformity achievable could be assessed. With the average dose across the profile normalized to 100%, hot spots up to 5.0% and troughs of 3% were identified for a composite beam of 2 x 5.0 mm grids, as measured at Dmax for a 6 MV beam. For a beam composed from 4 x 2.5 mm grids, the maximum across the profile was 3.0% above the average, and the minimum 2.5% below. Actual composite profiles were also formed using the integrating properties of film, with the subfield indexing performed using an engineering positioning stage. The beam uniformity for these fields compared well with that achieved in theory using the diode measurements. Finally sine wave patterns were generated to demonstrate the potential improvements in field shaping and conformity using this device as opposed to the conventional MLC alone. The scalloping effect on the field edge commonly seen on MLC fields was appreciably reduced by use of 2 x 5.0 mm

  18. Multileaf collimation of electrons - clinical effects on electron energy modulation and mixed beam therapy depending on treatment head design

    NASA Astrophysics Data System (ADS)

    Blomquist, Michael; Karlsson, Magnus G.; Zackrisson, Björn; Karlsson, Mikael

    2002-04-01

    The aim of this study was to explore the possibilities of using multileaf-collimated electron beams for advanced radiation therapy with conventional scattering foil flattened beams. Monte Carlo simulations were performed with the aim to improve electron beam characteristics and enable isocentric multileaf collimation. The scattering foil positions, monitor chamber thickness, the MLC location and the amount of He in the treatment head were optimized for three common commercial accelerators. The performance of the three optimized treatment head designs was compared for different SSDs in air, at treatment depth in water and for some clinical cases. The effects of electron/photon beam matching including generalized random and static errors using Gaussian one-dimensional (1D) error distributions, and also electron energy modulation, were studied at treatment depth in water. The modification of the treatment heads improved the electron beam characteristics and enabled the use of multileaf collimation in isocentric delivery of both electron and photon beams in a mixed beam IMRT procedure.

  19. Commissioning a Miniature Multileaf Collimator for Small Field Radiotherapy

    SciTech Connect

    Eaton, David Twyman, Nicola; Thomas, Simon

    2010-04-01

    The aim of this study was to determine dosimetric data for the Siemens ModuLeaf miniature multileaf collimator. Acquiring accurate data for small fields requires careful evaluation of different measurement techniques and dosimeters. Output factors were measured using several different detectors. Comparing these results demonstrated the relative under-response of the devices as the field size approaches the active area of the detector. Measurements obtained with a diamond detector, radiographic film, 0.125 cm{sup 3} cylindrical and advanced Markus ionization chambers agreed (within 0.7%) down to a field size of 2 x 2 cm. For smaller field sizes, only the diamond detector and the film agreed within their relative uncertainties. Head scatter factors were measured in brass and Perspex miniphantoms using the cylindrical chamber positioned at the isocenter, and in the Perspex miniphantom at extended distances. For field sizes of less than 3 x 3 cm, only the measurements with the brass miniphantom at the isocenter and the Perspex miniphantom at extended distances agreed (within 0.8%). The effect on the miniphantom measurements of using a low scatter support stand was also quantified. Beam profiles were measured using radiographic film. An unexpected artifact was found in these images because of increased backscatter from white lines inscribed and painted on the phantom. Percentage depth doses (PDDs) measured with and without the ModuLeaf attached agreed for all field sizes down to 3 x 3 cm. Below this field size, tissue maximum ratios (TMRs) were measured using an advanced Markus chamber. The 3 x 3 cm ModuLeaf TMR data agreed with the non-ModuLeaf PDD data within their relative uncertainties. This study investigated the use of several different measurement methods from which accurate dosimetric parameters can be obtained without the need to acquire specialist detectors.

  20. Pitfalls of tungsten multileaf collimator in proton beam therapy

    SciTech Connect

    Moskvin, Vadim; Cheng, Chee-Wai; Das, Indra J.

    2011-12-15

    Purpose: Particle beam therapy is associated with significant startup and operational cost. Multileaf collimator (MLC) provides an attractive option to improve the efficiency and reduce the treatment cost. A direct transfer of the MLC technology from external beam radiation therapy is intuitively straightforward to proton therapy. However, activation, neutron production, and the associated secondary cancer risk in proton beam should be an important consideration which is evaluated. Methods: Monte Carlo simulation with FLUKA particle transport code was applied in this study for a number of treatment models. The authors have performed a detailed study of the neutron generation, ambient dose equivalent [H*(10)], and activation of a typical tungsten MLC and compared with those obtained from a brass aperture used in a typical proton therapy system. Brass aperture and tungsten MLC were modeled by absorber blocks in this study, representing worst-case scenario of a fully closed collimator. Results: With a tungsten MLC, the secondary neutron dose to the patient is at least 1.5 times higher than that from a brass aperture. The H*(10) from a tungsten MLC at 10 cm downstream is about 22.3 mSv/Gy delivered to water phantom by noncollimated 200 MeV beam of 20 cm diameter compared to 14 mSv/Gy for the brass aperture. For a 30-fraction treatment course, the activity per unit volume in brass aperture reaches 5.3 x 10{sup 4} Bq cm{sup -3} at the end of the last treatment. The activity in brass decreases by a factor of 380 after 24 h, additional 6.2 times after 40 days of cooling, and is reduced to background level after 1 yr. Initial activity in tungsten after 30 days of treating 30 patients per day is about 3.4 times higher than in brass that decreases only by a factor of 2 after 40 days and accumulates to 1.2 x 10{sup 6} Bq cm{sup -3} after a full year of operation. The daily utilization of the MLC leads to buildup of activity with time. The overall activity continues to increase

  1. Monte Carlo simulations of electron beams collimated with a dual electron multileaf collimator: a feasibility study.

    PubMed

    Inyang, S O; Chamberlain, A C

    2009-07-01

    Electron applicators and cut-outs have been used for some time in the delivery of electron beam therapy. A dual electron multileaf collimator (eMLC) consisting of upper and lower eMLCs was designed, and dose distributions of electron beams delivered by it were evaluated by Monte Carlo simulations by use of the EGSnrc Monte Carlo code. The percentage depth dose (PDD), dose profiles, dose gradient falloff (G), depth of maximum dose (R(100)), surface dose, bremsstrahlung background, beam flatness, and penumbra of the dual eMLC were evaluated and compared with those simulated and measured with the standard applicators inserted into the treatment head of the medical linear accelerator (linac). The results showed good agreement in most cases. Specifically, the flatness and penumbra obtained with the dual eMLC were better than those obtained with standard applicators. It is therefore possible to use the proposed dual eMLC in the delivery of electron beam therapy without the need for applicators and cut-outs. This will minimize the inconvenience of placing cut-outs on the patients; also, changes in the required field can be effected without the therapist going into the accelerator room when the dual eMLC is finally automated. The absence of a helium bag between the upper and lower eMLCs did not offer significant changes in the dosimetry parameters compared to the eMLC fitted with the helium bag. PMID:20821122

  2. Measuring output factors of small fields formed by collimator jaws and multileaf collimator using plastic scintillation detectors

    SciTech Connect

    Klein, David M.; Tailor, Ramesh C.; Archambault, Louis; Wang, Lilie; Therriault-Proulx, Francois; Beddar, A. Sam

    2010-10-15

    Purpose: As the practice of using high-energy photon beams to create therapeutic radiation fields of subcentimeter dimensions (as in intensity-modulated radiotherapy or stereotactic radiosurgery) grows, so too does the need for accurate verification of beam output at these small fields in which standard practices of dose verification break down. This study investigates small-field output factors measured using a small plastic scintillation detector (PSD), as well as a 0.01 cm{sup 3} ionization chamber. Specifically, output factors were measured with both detectors using small fields that were defined by either the X-Y collimator jaws or the multileaf collimator (MLC). Methods: A PSD of 0.5 mm diameter and 2 mm length was irradiated with 6 and 18 MV linac beams. The PSD was positioned vertically at a source-to-axis distance of 100 cm, at 10 cm depth in a water phantom, and irradiated with fields ranging in size from 0.5x0.5 to 10x10 cm{sup 2}. The field sizes were defined either by the collimator jaws alone or by a MLC alone. The MLC fields were constructed in two ways: with the closed leaves (i.e., those leaves that were not opened to define the square field) meeting at either the field center line or at a 4 cm offset from the center line. Scintillation light was recorded using a CCD camera and an estimation of error in the median-filtered signals was made using the bootstrapping technique. Measurements were made using a CC01 ionization chamber under conditions identical to those used for the PSD. Results: Output factors measured by the PSD showed close agreement with those measured using the ionization chamber for field sizes of 2.0x2.0 cm{sup 2} and above. At smaller field sizes, the PSD obtained output factors as much as 15% higher than those found using the ionization chamber by 0.6x0.6 cm{sup 2} jaw-defined fields. Output factors measured with no offset of the closed MLC leaves were as much as 20% higher than those measured using a 4 cm leaf offset

  3. Lateral Penumbra Modelling Based Leaf End Shape Optimization for Multileaf Collimator in Radiotherapy

    PubMed Central

    Zhou, Dong; Zhang, Hui; Ye, Peiqing

    2016-01-01

    Lateral penumbra of multileaf collimator plays an important role in radiotherapy treatment planning. Growing evidence has revealed that, for a single-focused multileaf collimator, lateral penumbra width is leaf position dependent and largely attributed to the leaf end shape. In our study, an analytical method for leaf end induced lateral penumbra modelling is formulated using Tangent Secant Theory. Compared with Monte Carlo simulation and ray tracing algorithm, our model serves well the purpose of cost-efficient penumbra evaluation. Leaf ends represented in parametric forms of circular arc, elliptical arc, Bézier curve, and B-spline are implemented. With biobjective function of penumbra mean and variance introduced, genetic algorithm is carried out for approximating the Pareto frontier. Results show that for circular arc leaf end objective function is convex and convergence to optimal solution is guaranteed using gradient based iterative method. It is found that optimal leaf end in the shape of Bézier curve achieves minimal standard deviation, while using B-spline minimum of penumbra mean is obtained. For treatment modalities in clinical application, optimized leaf ends are in close agreement with actual shapes. Taken together, the method that we propose can provide insight into leaf end shape design of multileaf collimator. PMID:27110274

  4. Lateral Penumbra Modelling Based Leaf End Shape Optimization for Multileaf Collimator in Radiotherapy.

    PubMed

    Zhou, Dong; Zhang, Hui; Ye, Peiqing

    2016-01-01

    Lateral penumbra of multileaf collimator plays an important role in radiotherapy treatment planning. Growing evidence has revealed that, for a single-focused multileaf collimator, lateral penumbra width is leaf position dependent and largely attributed to the leaf end shape. In our study, an analytical method for leaf end induced lateral penumbra modelling is formulated using Tangent Secant Theory. Compared with Monte Carlo simulation and ray tracing algorithm, our model serves well the purpose of cost-efficient penumbra evaluation. Leaf ends represented in parametric forms of circular arc, elliptical arc, Bézier curve, and B-spline are implemented. With biobjective function of penumbra mean and variance introduced, genetic algorithm is carried out for approximating the Pareto frontier. Results show that for circular arc leaf end objective function is convex and convergence to optimal solution is guaranteed using gradient based iterative method. It is found that optimal leaf end in the shape of Bézier curve achieves minimal standard deviation, while using B-spline minimum of penumbra mean is obtained. For treatment modalities in clinical application, optimized leaf ends are in close agreement with actual shapes. Taken together, the method that we propose can provide insight into leaf end shape design of multileaf collimator. PMID:27110274

  5. Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration.

    PubMed

    Baker, S J K; Budgell, G J; MacKay, R I

    2005-04-01

    Multileaf collimator (MLC) calibration and quality control is a time-consuming procedure typically involving the processing, scanning and analysis of films to measure leaf and collimator positions. Faster and more reliable calibration procedures are required for these tasks, especially with the introduction of intensity modulated radiotherapy which requires more frequent checking and finer positional leaf tolerances than previously. A routine quality control (QC) technique to measure MLC leaf bank gain and offset, as well as minor offsets (individual leaf position relative to a reference leaf), using an amorphous silicon electronic portal imaging device (EPID) has been developed. The technique also tests the calibration of the primary and back-up collimators. A detailed comparison between film and EPID measurements has been performed for six linear accelerators (linacs) equipped with MLC and amorphous silicon EPIDs. Measurements of field size from 4 to 24 cm with the EPID were systematically smaller than film measurements over all field sizes by 0.4 mm for leaves/back-up collimators and by 0.2 mm for conventional collimators. This effect is due to the gain calibration correction applied by the EPID, resulting in a 'flattening' of primary beam profiles. Linac dependent systematic differences of up to 0.5 mm in individual leaf/collimator positions were also found between EPID and film measurements due to the difference between the mechanical and radiation axes of rotation. When corrections for these systematic differences were applied, the residual random differences between EPID and film were 0.23 mm and 0.26 mm (1 standard deviation) for field size and individual leaf/back-up collimator position, respectively. Measured gains (over a distance of 220 mm) always agreed within 0.4 mm with a standard deviation of 0.17 mm. Minor offset measurements gave a mean agreement between EPID and film of 0.01+/-0.10 mm (1 standard deviation) after correction for the tilt of the

  6. Compact multileaf collimator for conformal and intensity modulated fast neutron therapy: electromechanical design and validation.

    PubMed

    Farr, J B; Maughan, R L; Yudelev, M; Blosser, E; Brandon, J; Horste, T; Forman, J D

    2006-09-01

    The electromechanical properties of a 120-leaf, high-resolution, computer-controlled, fast neutron multileaf collimator (MLC) are presented. The MLC replaces an aging, manually operated multirod collimator. The MLC leaves project 5 mm in the isocentric plane perpendicular to the beam axis. A taper is included on the leaves matching beam divergence along one axis. The 5-mm leaf projection width is chosen to give high-resolution conformality across the entire field. The maximum field size provided is 30 x 30 cm2. To reduce the interleaf transmission a 0.254-mm blocking step is included. End-leaf steps totaling 0.762 mm are also provided allowing opposing leaves to close off within the primary radiation beam. The neutron MLC also includes individual 45 degrees and 60 degrees automated universal tungsten wedges. The automated high-resolution neutron collimation provides an increase in patient throughput capacity, enables a new modality, intensity modulated neutron therapy, and limits occupational radiation exposure by providing remote operation from a shielded console area. PMID:17022226

  7. SU-E-T-247: Multi-Leaf Collimator Model Adjustments Improve Small Field Dosimetry in VMAT Plans

    SciTech Connect

    Young, L; Yang, F

    2014-06-01

    Purpose: The Elekta beam modulator linac employs a 4-mm micro multileaf collimator (MLC) backed by a fixed jaw. Out-of-field dose discrepancies between treatment planning system (TPS) calculations and output water phantom measurements are caused by the 1-mm leaf gap required for all moving MLCs in a VMAT arc. In this study, MLC parameters are optimized to improve TPS out-of-field dose approximations. Methods: Static 2.4 cm square fields were created with a 1-mm leaf gap for MLCs that would normally park behind the jaw. Doses in the open field and leaf gap were measured with an A16 micro ion chamber and EDR2 film for comparison with corresponding point doses in the Pinnacle TPS. The MLC offset table and tip radius were adjusted until TPS point doses agreed with photon measurements. Improvements to the beam models were tested using static arcs consisting of square fields ranging from 1.6 to 14.0 cm, with 45° collimator rotation, and 1-mm leaf gap to replicate VMAT conditions. Gamma values for the 3-mm distance, 3% dose difference criteria were evaluated using standard QA procedures with a cylindrical detector array. Results: The best agreement in point doses within the leaf gap and open field was achieved by offsetting the default rounded leaf end table by 0.1 cm and adjusting the leaf tip radius to 13 cm. Improvements in TPS models for 6 and 10 MV photon beams were more significant for smaller field sizes 3.6 cm or less where the initial gamma factors progressively increased as field size decreased, i.e. for a 1.6cm field size, the Gamma increased from 56.1% to 98.8%. Conclusion: The MLC optimization techniques developed will achieve greater dosimetric accuracy in small field VMAT treatment plans for fixed jaw linear accelerators. Accurate predictions of dose to organs at risk may reduce adverse effects of radiotherapy.

  8. Characterization of a mini-multileaf collimator in a proton beamline

    SciTech Connect

    Daartz, J.; Bangert, M.; Bussiere, M. R.; Engelsman, M.; Kooy, H. M.

    2009-05-15

    A mini-multileaf collimator (MMLC) was mounted as a field shaping collimator in a proton beamline at the Massachusetts General Hospital. The purpose is to evaluate the device's dosimetric and mechanical properties for the use in a proton beamline. For this evaluation, the authors compared MMLC and brass aperture shaped dose distributions with regard to lateral and depth dose properties. The lateral fall off is generally broader with the MMLC, with difference varying with proton range from 0.2 to 1.2 mm. Central axis depth dose curves did not show a difference in peak-to-entrance ratio, peak width, distal fall off, or range. Two-dimensional dose distributions to investigate the conformity of MMLC shaped doses show that the physical leaf width of {approx_equal}2.5 mm does not have a significant impact. All differences seen in dose distribution shaped by the MMLC versus brass apertures were shown to be clinically insignificant. Measured neutron doses of 0.03-0.13 mSv/Gy for a closed brass beam block (depending on range) are very low compared to the previously published data. Irradiation of the tungsten MMLC, however, produced 1.5-1.8 times more neutrons than brass apertures. Exposure of the staff resulting from activation of the device is below regulatory limits. The measurements established an equivalency between aperture and MMLC shaped dose distributions.

  9. Multibeam tomotherapy: a new treatment unit devised for multileaf collimation, intensity-modulated radiation therapy.

    PubMed

    Achterberg, Nils; Müller, Reinhold G

    2007-10-01

    A fully integrated system for treatment planning, application, and verification for automated multileaf collimator (MLC) based, intensity-modulated, image-guided, and adaptive radiation therapy (IMRT, IGRT and ART, respectively) is proposed. Patient comfort, which was the major development goal, will be achieved through a new unit design and short treatment times. Our device for photon beam therapy will consist of a new dual energy linac with five fixed treatment heads positioned evenly along one plane but one electron beam generator only. A minimum of moving parts increases technical reliability and reduces motion times to a minimum. Motion is allowed solely for the MLCs, the robotic patient table, and the small angle gantry rotation of +/- 36 degrees. Besides sophisticated electron beam guidance, this compact setup can be built using existing modules. The flattening-filter-free treatment heads are characterized by reduced beam-on time and contain apertures restricted in one dimension to the area of maximum primary fluence output. In the case of longer targets, this leads to a topographic intensity modulation, thanks to the combination of "step and shoot" MLC delivery and discrete patient couch motion. Owing to the limited number of beam directions, this multislice cone beam serial tomotherapy is referred to as "multibeam tomotherapy." Every patient slice is irradiated by one treatment head at any given moment but for one subfield only. The electron beam is then guided to the next head ready for delivery, while the other heads are preparing their leaves for the next segment. The "Multifocal MLC-positioning" algorithm was programmed to enable treatment planning and optimize treatment time. We developed an overlap strategy for the longitudinally adjacent fields of every beam direction, in doing so minimizing the field match problem and the effects of possible table step errors. Clinical case studies show for the same or better planning target volume coverage, better

  10. Multibeam tomotherapy: A new treatment unit devised for multileaf collimation, intensity-modulated radiation therapy

    SciTech Connect

    Achterberg, Nils; Mueller, Reinhold G.

    2007-10-15

    A fully integrated system for treatment planning, application, and verification for automated multileaf collimator (MLC) based, intensity-modulated, image-guided, and adaptive radiation therapy (IMRT, IGRT and ART, respectively) is proposed. Patient comfort, which was the major development goal, will be achieved through a new unit design and short treatment times. Our device for photon beam therapy will consist of a new dual energy linac with five fixed treatment heads positioned evenly along one plane but one electron beam generator only. A minimum of moving parts increases technical reliability and reduces motion times to a minimum. Motion is allowed solely for the MLCs, the robotic patient table, and the small angle gantry rotation of {+-}36 deg. . Besides sophisticated electron beam guidance, this compact setup can be built using existing modules. The flattening-filter-free treatment heads are characterized by reduced beam-on time and contain apertures restricted in one dimension to the area of maximum primary fluence output. In the case of longer targets, this leads to a topographic intensity modulation, thanks to the combination of 'step and shoot' MLC delivery and discrete patient couch motion. Owing to the limited number of beam directions, this multislice cone beam serial tomotherapy is referred to as 'multibeam tomotherapy.' Every patient slice is irradiated by one treatment head at any given moment but for one subfield only. The electron beam is then guided to the next head ready for delivery, while the other heads are preparing their leaves for the next segment. The 'Multifocal MLC-positioning' algorithm was programmed to enable treatment planning and optimize treatment time. We developed an overlap strategy for the longitudinally adjacent fields of every beam direction, in doing so minimizing the field match problem and the effects of possible table step errors. Clinical case studies show for the same or better planning target volume coverage, better

  11. Assessment of a new multileaf collimator concept using GEANT4 Monte Carlo simulations.

    PubMed

    Tacke, Martin B; Szymanowski, Hanitra; Oelfke, Uwe; Schulze, Carsten; Nuss, Susanne; Wehrwein, Eugen; Leidenberger, Stefan

    2006-04-01

    The aim of the work was to investigate in advance the dosimetric properties of a new multileaf collimator (MLC) concept with the help of Monte Carlo (MC) simulations prior to the production of a prototype. The geometrical design of the MLC was implemented in the MC code GEANT4. For the simulation of a 6 MV treatment beam, an experimentally validated phase space and a virtual spatial Gaussian-shaped model placed in the origin were used. For the simulation of the geometry in GEANT4, the jaws and the two leaf packages were implemented with the help of computer-aided design data. First, transmission values for different tungsten alloys were extracted using the simulation codes GEANT4 and BEAMnrc and compared to experimental measurements. In a second step, high-resolution simulations were performed to detect the leakage at depth of maximum dose. The 20%-80% penumbra along the travel direction of the leaves was determined using 10 x 10 cm2 fields shifted along the x- and y-axis. The simulated results were compared with measured data. The simulation of the transmission values for different tungsten alloys showed a good agreement with the experimental measurements (within 2.0%). This enabled an accurate estimation of the attenuation coefficient for the various leaf materials. Simulations with varying width of the spatial Gaussian distribution showed that the leakage and the penumbra depend very much on this parameter: for instance, for widths of 2 and 4 mm, the interleaf leakage is below 0.3% and 0.75%, respectively. The results for the leakage and the penumbra (4.7+/-0.5 mm) are in good agreement with the measurements. This study showed that GEANT4 is appropriate for the investigation of the dosimetric properties of a multileaf collimator. In particular, a quantification of the leakage, the penumbra, and the tongue-and-groove effect and an evaluation of the influence of the beam parameters such as the width of the Gaussian distribution was possible.

  12. Evaluation of dosimetric characteristics of multi-leaf and conventional collimated radiation fields using a scanning liquid ionization chamber EPID.

    PubMed

    Mohammadi, M; Bezak, E

    2008-12-01

    The characteristics of radiation fields set up using conventional and Multi-Leaf collimators were investigated using a Scanning Liquid Ionization Chamber Electronic Portal Imaging Device (SLIC-EPID). Results showed that the radiation fields set up using MLCs are generally larger than those set up using conventional collimators. A significant difference was observed between the penumbra width for conventional and MLC radiation fields. SLIC-EPID was found to be a sensitive device to evaluate the characteristics of the radiation fields generated with MLCs.

  13. An independent system for real-time dynamic multileaf collimation trajectory verification using EPID

    NASA Astrophysics Data System (ADS)

    Fuangrod, Todsaporn; Woodruff, Henry C.; Rowshanfarzad, Pejman; O'Connor, Daryl J.; Middleton, Richard H.; Greer, Peter B.

    2014-01-01

    A new tool has been developed to verify the trajectory of dynamic multileaf collimators (MLCs) used in advanced radiotherapy techniques using only the information provided by the electronic portal imaging devices (EPID) measured image frames. The prescribed leaf positions are resampled to a higher resolution in a pre-processing stage to improve the verification precision. Measured MLC positions are extracted from the EPID frames using a template matching method. A cosine similarity metric is then applied to synchronise measured and planned leaf positions for comparison. Three additional comparison functions were incorporated to ensure robust synchronisation. The MLC leaf trajectory error detection was simulated for both intensity modulated radiation therapy (IMRT) (prostate) and volumetric modulated arc therapy (VMAT) (head-and-neck) deliveries with anthropomorphic phantoms in the beam. The overall accuracy for MLC positions automatically extracted from EPID image frames was approximately 0.5 mm. The MLC leaf trajectory verification system can detect leaf position errors during IMRT and VMAT with a tolerance of 3.5 mm within 1 s.

  14. Design and Fabrication of the Control Part of a Prototype Multileaf Collimator System

    PubMed Central

    Hashemian, Abdolreza; Toossi, Mohammad Taghi Bahreyni; Nasseri, Shahrokh

    2014-01-01

    Multileaf collimator (MLC) is among the radiation field shaping systems used for conformal radiotherapy and intensity modulation radiation therapy techniques. The MLC system that has been designed and fabricated in this study includes 52 leaves, 52 stepper motors, 2 DC motors, 16 programmable logic controllers (PLCs) and one human machine interface (HMI). This system can be mounted on conventional linear accelerators (linac) as an add-on accessory. The 52 leaves are mounted on two carriages that are moved independently. The leaves sequence acquired from the image processing of computed tomography images is used to arrange leaves. This sequence is saved in a text file. The leaves are arranged by HMI and labVIEW. Using HMI it is possible to test the operation of PLCs and manually enter the numerical values of the leaves edges. An executable file is developed by labVIEW program, which is graphically user interfaced between the operator and the MLC control system. The projected width of each leaf on the isocenter accelerator (usually at 100 cm from the source) is 10 mm. The positioning accuracy of each leaf is approximately 1.4 mm. PMID:25426434

  15. Postoperative modified stereotactic radiotherapy using a micro-multileaf collimator in patients with malignant glioma.

    PubMed

    Isaka, Toshihiko; Nishiyama, Kinji; Nakagawa, Hidemitsu; Suzuki, Tsuyoshi; Wada, Kouichi

    2002-06-01

    To achieve local control of malignant glioma, we designed a postoperative stereotactic radiotherapy using a micro-multileaf collimator (micro-MLC). The purpose of this study was to clarify the feasibility of this treatment. The treatment was performed in six patients who met the following eligibility criteria: (1) supratentorial tumor, (2) residual tumor volume < or = 40 cm3, and (3) Karnofsky performance status > or = 70. The three planning target volumes (PTVs), which consisted of restricted PTV (RPTV), intermediate PTV (IPTV), and extended PTV (EPTV), defined as the residual tumor plus a 1 cm, 2 cm, and 3 cm margins, respectively, and total dose delivery of 60-68 Gy, 52-60 Gy, and 44-52 Gy to the isocenters of RPTV, IPTV, and EPTV, respectively, in 4 Gy per fraction at five fractions per week, were established. The beam arrangement and the conformal blockade with a micro-MLC for the optimal treatment plan were designed. The treatment plans showed the high dose conformation to EPTV, the appropriate dose gradients in the three PTVs with the high dose homogeneity to RPTV, and the tolerated dose to critical structures. Following the plans, treatment was performed. The clinical findings more than 12 months after the treatment supported its possible use. We conclude that this treatment is feasible at least in selected patients.

  16. Variable Circular Collimator in Robotic Radiosurgery: A Time-Efficient Alternative to a Mini-Multileaf Collimator?

    SciTech Connect

    Water, Steven van de; Hoogeman, Mischa S.; Breedveld, Sebastiaan; Nuyttens, Joost J.M.E.; Schaart, Dennis R.; Heijmen, Ben J.M.

    2011-11-01

    Purpose: Compared with many small circular beams used in CyberKnife treatments, beam's eye view-shaped fields are generally more time-efficient for dose delivery. However, beam's eye view-shaping devices, such as a mini-multileaf collimator (mMLC), are not presently available for CyberKnife, although a variable-aperture collimator (Iris, 12 field diameters; 5-60 mm) is available. We investigated whether the Iris can mimic noncoplanar mMLC treatments using a limited set of principal beam orientations (nodes) to produce time-efficient treatment plans. Methods and Materials: The data from 10 lung cancer patients and the beam-orientation optimization algorithm 'Cycle' were used to generate stereotactic treatment plans (3 x 20 Gy) for a CyberKnife virtually equipped with a mMLC. Typically, 10-16 favorable beam orientations were selected from 117 available robot node positions using beam's eye view-shaped fields with uniform fluence. Second, intensity-modulated Iris plans were generated by inverse optimization of nonisocentric circular candidate beams targeted from the same nodes selected in the mMLC plans. The plans were evaluated using the mean lung dose, lung volume receiving {>=}20 Gy, conformality index, number of nodes, beams, and monitor units, and estimated treatment time. Results: The mMLC plans contained an average of 12 nodes and 11,690 monitor units. For a comparable mean lung dose, the Iris plans contained 12 nodes, 64 beams, and 21,990 monitor units. The estimated fraction duration was 12.2 min (range, 10.8-13.5) for the mMLC plans and 18.4 min (range, 12.9-28.5) for the Iris plans. In contrast to the mMLC plans, the treatment time for the Iris plans increased with an increasing target volume. The Iris plans were, on average, 40% longer than the corresponding mMLC plans for small targets (<80 cm{sup 3}) and {<=}121% longer for larger targets. For a comparable conformality index, similar results were obtained. Conclusion: For stereotactic lung irradiation

  17. Mantle fields in the era of dynamic multileaf collimation: Field shaping and electronic tissue compensation

    SciTech Connect

    Davis, Quillin G.; Paulino, Arnold C. . E-mail: apaulino@tmh.tmc.edu; Miller, Robin M.S.; Ting, Joseph Y.

    2006-10-01

    Mantle field radiotherapy for Hodgkin's disease is complicated by significant dose gradient (up to 10-20%) across the large fields required. Many different strategies of tissue compensation have been investigated, including custom physical compensators to provide better dose distributions. We present a method using dynamic multileaf collimator (dMLC) fluence modulation to simultaneously shape the treatment field and give homogeneous dose at depth throughout the classic mantle field. Five patients were treated for early-stage Hodgkin's disease with a conventional anterior-posterior-posterior-anterior (AP-PA) mantle field. The patients were planned using the Varian Eclipse treatment planning system, version 6.1.3, and treated on a Varian 2300CD. An AP-PA dynamic MLC beam-shaped and dose-compensated plan was created for each, and compared with the conventional blocked plan. Nine dose points were calculated at midplane in each plan. Chamber measurements were taken to confirm accurate dMLC delivery of the planned doses. The mean dose per fraction, relative to a central axis dose of 1.8 Gy, was increased in the conventional plans compared with the dMLC plans in the right (R) neck, left (L) neck, R supraclavicular, L supraclavicular, and L axillary points. The mediastinum tended to be underdosed relative to central axis, with the mid-mediastinal and lower mediastinal points showing improved coverage with the dMLC plans. Measurements showed excellent agreement between planned doses and delivered doses, with less than 2% in-field variation. Dynamic MLC fluence modulation was used to effectively deliver a mantle field that is both shape- and electronically-dose-compensated with sliding window MLC. Homogeneity was significantly improved throughout the treatment field, and measurements confirmed accurate dose delivery using this technique.

  18. A machine learning approach to the accurate prediction of multi-leaf collimator positional errors

    NASA Astrophysics Data System (ADS)

    Carlson, Joel N. K.; Park, Jong Min; Park, So-Yeon; In Park, Jong; Choi, Yunseok; Ye, Sung-Joon

    2016-03-01

    Discrepancies between planned and delivered movements of multi-leaf collimators (MLCs) are an important source of errors in dose distributions during radiotherapy. In this work we used machine learning techniques to train models to predict these discrepancies, assessed the accuracy of the model predictions, and examined the impact these errors have on quality assurance (QA) procedures and dosimetry. Predictive leaf motion parameters for the models were calculated from the plan files, such as leaf position and velocity, whether the leaf was moving towards or away from the isocenter of the MLC, and many others. Differences in positions between synchronized DICOM-RT planning files and DynaLog files reported during QA delivery were used as a target response for training of the models. The final model is capable of predicting MLC positions during delivery to a high degree of accuracy. For moving MLC leaves, predicted positions were shown to be significantly closer to delivered positions than were planned positions. By incorporating predicted positions into dose calculations in the TPS, increases were shown in gamma passing rates against measured dose distributions recorded during QA delivery. For instance, head and neck plans with 1%/2 mm gamma criteria had an average increase in passing rate of 4.17% (SD  =  1.54%). This indicates that the inclusion of predictions during dose calculation leads to a more realistic representation of plan delivery. To assess impact on the patient, dose volumetric histograms (DVH) using delivered positions were calculated for comparison with planned and predicted DVHs. In all cases, predicted dose volumetric parameters were in closer agreement to the delivered parameters than were the planned parameters, particularly for organs at risk on the periphery of the treatment area. By incorporating the predicted positions into the TPS, the treatment planner is given a more realistic view of the dose distribution as it will truly be

  19. Monte Carlo simulation based study of a proposed multileaf collimator for a telecobalt machine

    SciTech Connect

    Sahani, G.; Dash Sharma, P. K.; Hussain, S. A.; Dutt Sharma, Sunil; Sharma, D. N.

    2013-02-15

    Purpose: The objective of the present work was to propose a design of a secondary multileaf collimator (MLC) for a telecobalt machine and optimize its design features through Monte Carlo simulation. Methods: The proposed MLC design consists of 72 leaves (36 leaf pairs) with additional jaws perpendicular to leaf motion having the capability of shaping a maximum square field size of 35 Multiplication-Sign 35 cm{sup 2}. The projected widths at isocenter of each of the central 34 leaf pairs and 2 peripheral leaf pairs are 10 and 5 mm, respectively. The ends of the leaves and the x-jaws were optimized to obtain acceptable values of dosimetric and leakage parameters. Monte Carlo N-Particle code was used for generating beam profiles and depth dose curves and estimating the leakage radiation through the MLC. A water phantom of dimension 50 Multiplication-Sign 50 Multiplication-Sign 40 cm{sup 3} with an array of voxels (4 Multiplication-Sign 0.3 Multiplication-Sign 0.6 cm{sup 3}= 0.72 cm{sup 3}) was used for the study of dosimetric and leakage characteristics of the MLC. Output files generated for beam profiles were exported to the PTW radiation field analyzer software through locally developed software for analysis of beam profiles in order to evaluate radiation field width, beam flatness, symmetry, and beam penumbra. Results: The optimized version of the MLC can define radiation fields of up to 35 Multiplication-Sign 35 cm{sup 2} within the prescribed tolerance values of 2 mm. The flatness and symmetry were found to be well within the acceptable tolerance value of 3%. The penumbra for a 10 Multiplication-Sign 10 cm{sup 2} field size is 10.7 mm which is less than the generally acceptable value of 12 mm for a telecobalt machine. The maximum and average radiation leakage through the MLC were found to be 0.74% and 0.41% which are well below the International Electrotechnical Commission recommended tolerance values of 2% and 0.75%, respectively. The maximum leakage through the

  20. Markerless EPID image guided dynamic multi-leaf collimator tracking for lung tumors.

    PubMed

    Rottmann, J; Keall, P; Berbeco, R

    2013-06-21

    Compensation of target motion during the delivery of radiotherapy has the potential to improve treatment accuracy, dose conformity and sparing of healthy tissue. We implement an online image guided therapy system based on soft tissue localization (STiL) of the target from electronic portal images and treatment aperture adaptation with a dynamic multi-leaf collimator (DMLC). The treatment aperture is moved synchronously and in real time with the tumor during the entire breathing cycle. The system is implemented and tested on a Varian TX clinical linear accelerator featuring an AS-1000 electronic portal imaging device (EPID) acquiring images at a frame rate of 12.86 Hz throughout the treatment. A position update cycle for the treatment aperture consists of four steps: in the first step at time t = t0 a frame is grabbed, in the second step the frame is processed with the STiL algorithm to get the tumor position at t = t0, in a third step the tumor position at t = ti + δt is predicted to overcome system latencies and in the fourth step, the DMLC control software calculates the required leaf motions and applies them at time t = ti + δt. The prediction model is trained before the start of the treatment with data representing the tumor motion. We analyze the system latency with a dynamic chest phantom (4D motion phantom, Washington University). We estimate the average planar position deviation between target and treatment aperture in a clinical setting by driving the phantom with several lung tumor trajectories (recorded from fiducial tracking during radiotherapy delivery to the lung). DMLC tracking for lung stereotactic body radiation therapy without fiducial markers was successfully demonstrated. The inherent system latency is found to be δt = (230 ± 11) ms for a MV portal image acquisition frame rate of 12.86 Hz. The root mean square deviation between tumor and aperture position is smaller than 1 mm. We demonstrate the feasibility of real-time markerless DMLC

  1. 6 MV dosimetric characterization of the 160 MLC, the new Siemens multileaf collimator

    SciTech Connect

    Tacke, Martin B.; Nill, Simeon; Haering, Peter; Oelfke, Uwe

    2008-05-15

    New technical developments constantly aim at improving the outcome of radiation therapy. With the use of a computer-controlled multileaf collimator (MLC), the quality of the treatment and the efficiency in patient throughput is significantly increased. New MLC designs aim to further enhance the advantages. In this article, we present the first detailed experimental investigation of the new 160 MLC{sup TM}, Siemens Medical Solutions. The assessment included the experimental investigation of typical MLC characteristics such as leakage, tongue-and-groove effect, penumbra, leaf speed, and leaf positioning accuracy with a 6 MV treatment beam. The leakage is remarkably low with an average of 0.37% due to a new design principle of slightly tilted leaves instead of the common tongue-and-groove design. But due to the tilt, the triangular tongue-and-groove effect occurs. Its magnitude of approximately 19% is similar to the dose defect measured for MLCs with the common tongue-and-groove design. The average longitudinal penumbra measured at depth d{sub max}=15 mm with standard 100x100 mm{sup 2} fields is 4.1{+-}0.5 mm for the central range and increases to 4.9{+-}1.3 mm for the entire field range of 400x400 mm{sup 2}. The increase is partly due to the single-focusing design and the large distance between the MLC and the isocenter enabling a large patient clearance. Regarding the leaf speed, different velocity tests were performed. The positions of the moving leaves were continuously recorded with the kilovoltage-imaging panel. The maximum leaf velocities measured were 42.9{+-}0.6 mm/s. In addition, several typical intensity-modulated radiation therapy treatments were performed and the delivery times compared to the Siemens OPTIFOCUS MLC. An average decrease of 11% in delivery time was observed. The experimental results presented in this article indicate that the dosimetric characteristics of the 160 MLC are capable of improving the quality of dose delivery with respect to

  2. Markerless EPID image guided dynamic multi-leaf collimator tracking for lung tumors.

    PubMed

    Rottmann, J; Keall, P; Berbeco, R

    2013-06-21

    Compensation of target motion during the delivery of radiotherapy has the potential to improve treatment accuracy, dose conformity and sparing of healthy tissue. We implement an online image guided therapy system based on soft tissue localization (STiL) of the target from electronic portal images and treatment aperture adaptation with a dynamic multi-leaf collimator (DMLC). The treatment aperture is moved synchronously and in real time with the tumor during the entire breathing cycle. The system is implemented and tested on a Varian TX clinical linear accelerator featuring an AS-1000 electronic portal imaging device (EPID) acquiring images at a frame rate of 12.86 Hz throughout the treatment. A position update cycle for the treatment aperture consists of four steps: in the first step at time t = t0 a frame is grabbed, in the second step the frame is processed with the STiL algorithm to get the tumor position at t = t0, in a third step the tumor position at t = ti + δt is predicted to overcome system latencies and in the fourth step, the DMLC control software calculates the required leaf motions and applies them at time t = ti + δt. The prediction model is trained before the start of the treatment with data representing the tumor motion. We analyze the system latency with a dynamic chest phantom (4D motion phantom, Washington University). We estimate the average planar position deviation between target and treatment aperture in a clinical setting by driving the phantom with several lung tumor trajectories (recorded from fiducial tracking during radiotherapy delivery to the lung). DMLC tracking for lung stereotactic body radiation therapy without fiducial markers was successfully demonstrated. The inherent system latency is found to be δt = (230 ± 11) ms for a MV portal image acquisition frame rate of 12.86 Hz. The root mean square deviation between tumor and aperture position is smaller than 1 mm. We demonstrate the feasibility of real-time markerless DMLC

  3. Dosimetric performance of the new high-definition multileaf collimator for intracranial stereotactic radiosurgery.

    PubMed

    Dhabaan, Anees; Elder, Eric; Schreibmann, Eduard; Crocker, Ian; Curran, Walter J; Oyesiku, Nelson M; Shu, Hui-Kuo; Fox, Tim

    2010-06-21

    The objective was to evaluate the performance of a high-definition multileaf collimator (MLC) of 2.5 mm leaf width (MLC2.5) and compare to standard 5 mm leaf width MLC (MLC5) for the treatment of intracranial lesions using dynamic conformal arcs (DCA) technique with a dedicated radiosurgery linear accelerator. Simulated cases of spherical targets were created to study solely the effect of target volume size on the performance of the two MLC systems independent of target shape complexity. In addition, 43 patients previously treated for intracranial lesions in our institution were retrospectively planned using DCA technique with MLC2.5 and MLC5 systems. The gross tumor volume ranged from 0.07 to 40.57 cm3 with an average volume of 5.9 cm3. All treatment parameters were kept the same for both MLC-based plans. The plan evaluation was performed using figures of merits (FOM) for a rapid and objective assessment on the quality of the two treatment plans for MLC2.5 and MLC5. The prescription isodose surface was selected as the greatest isodose surface covering >or= 95% of the target volume and delivering 95% of the prescription dose to 99% of target volume. A Conformity Index (CI) and conformity distance index (CDI) were used to quantifying the dose conformity to a target volume. To assess normal tissue sparing, a normal tissue difference (NTD) was defined as the difference between the volume of normal tissue receiving a certain dose utilizing MLC5 and the volume receiving the same dose using MLC2.5. The CI and normal tissue sparing for the simulated spherical targets were better with the MLC2.5 as compared to MLC5. For the clinical patients, the CI and CDI results indicated that the MLC2.5 provides better treatment conformity than MLC5 even at large target volumes. The CI's range was 1.15 to 2.44 with a median of 1.59 for MLC2.5 compared to 1.60-2.85 with a median of 1.71 for MLC5. Improved normal tissue sparing was also observed for MLC2.5 over MLC5, with the NTD always

  4. Markerless EPID image guided dynamic multi-leaf collimator tracking for lung tumors

    NASA Astrophysics Data System (ADS)

    Rottmann, J.; Keall, P.; Berbeco, R.

    2013-06-01

    Compensation of target motion during the delivery of radiotherapy has the potential to improve treatment accuracy, dose conformity and sparing of healthy tissue. We implement an online image guided therapy system based on soft tissue localization (STiL) of the target from electronic portal images and treatment aperture adaptation with a dynamic multi-leaf collimator (DMLC). The treatment aperture is moved synchronously and in real time with the tumor during the entire breathing cycle. The system is implemented and tested on a Varian TX clinical linear accelerator featuring an AS-1000 electronic portal imaging device (EPID) acquiring images at a frame rate of 12.86 Hz throughout the treatment. A position update cycle for the treatment aperture consists of four steps: in the first step at time t = t0 a frame is grabbed, in the second step the frame is processed with the STiL algorithm to get the tumor position at t = t0, in a third step the tumor position at t = ti + δt is predicted to overcome system latencies and in the fourth step, the DMLC control software calculates the required leaf motions and applies them at time t = ti + δt. The prediction model is trained before the start of the treatment with data representing the tumor motion. We analyze the system latency with a dynamic chest phantom (4D motion phantom, Washington University). We estimate the average planar position deviation between target and treatment aperture in a clinical setting by driving the phantom with several lung tumor trajectories (recorded from fiducial tracking during radiotherapy delivery to the lung). DMLC tracking for lung stereotactic body radiation therapy without fiducial markers was successfully demonstrated. The inherent system latency is found to be δt = (230 ± 11) ms for a MV portal image acquisition frame rate of 12.86 Hz. The root mean square deviation between tumor and aperture position is smaller than 1 mm. We demonstrate the feasibility of real-time markerless DMLC

  5. Collimator-free photon tomography

    DOEpatents

    Dilmanian, F. Avraham; Barbour, Randall L.

    1998-10-06

    A method of uncollimated single photon emission computed tomography includes administering a radioisotope to a patient for producing gamma ray photons from a source inside the patient. Emissivity of the photons is measured externally of the patient with an uncollimated gamma camera at a plurality of measurement positions surrounding the patient for obtaining corresponding energy spectrums thereat. Photon emissivity at the plurality of measurement positions is predicted using an initial prediction of an image of the source. The predicted and measured photon emissivities are compared to obtain differences therebetween. Prediction and comparison is iterated by updating the image prediction until the differences are below a threshold for obtaining a final prediction of the source image.

  6. Collimator-free photon tomography

    DOEpatents

    Dilmanian, F.A.; Barbour, R.L.

    1998-10-06

    A method of uncollimated single photon emission computed tomography includes administering a radioisotope to a patient for producing gamma ray photons from a source inside the patient. Emissivity of the photons is measured externally of the patient with an uncollimated gamma camera at a plurality of measurement positions surrounding the patient for obtaining corresponding energy spectrums thereat. Photon emissivity at the plurality of measurement positions is predicted using an initial prediction of an image of the source. The predicted and measured photon emissivities are compared to obtain differences therebetween. Prediction and comparison is iterated by updating the image prediction until the differences are below a threshold for obtaining a final prediction of the source image. 6 figs.

  7. Super-collimation by axisymmetric photonic crystals

    SciTech Connect

    Purlys, V.; Gailevičius, D.; Peckus, M.; Gadonas, R.; Maigyte, L.; Staliunas, K.

    2014-06-02

    We propose and experimentally show the mechanism of beam super-collimation by axisymmetric photonic crystals, specifically by periodic (in propagation direction) structure of layers of concentric rings. The physical mechanism behind the effect is an inverse scattering cascade of diffracted wave components back into on- and near-axis angular field components, resulting in substantial enhancement of intensity of these components. We explore the super-collimation by numerical calculations and prove it experimentally. We demonstrate experimentally the axial field enhancement up to 7 times in terms of field intensity.

  8. Photonic crystal self-collimation sensor.

    PubMed

    Wang, Yufei; Wang, Hailing; Xue, Qikun; Zheng, Wanhua

    2012-05-21

    A novel refractive index sensor based on the two dimensional photonic crystal folded Michelson interferometer employing the self-collimation effect is proposed and its performances are theoretically investigated. Two sensing areas are included in the sensor. Simulation results indicate the branch area is suitable for the small index variety range and fine detection, whereas the reflector area prone to the large index change range and coarse detection. Because of no defect waveguides and no crosstalk of signal, the sensor is desirable to perform monolithic integrated, low-cost, label-free real-time parallel sensing. In addition, a flexible design of self-collimation sensors array is demonstrated. PMID:22714197

  9. Multileaf collimator performance monitoring and improvement using semiautomated quality control testing and statistical process control

    SciTech Connect

    Létourneau, Daniel McNiven, Andrea; Keller, Harald; Wang, An; Amin, Md Nurul; Pearce, Jim; Norrlinger, Bernhard; Jaffray, David A.

    2014-12-15

    Purpose: High-quality radiation therapy using highly conformal dose distributions and image-guided techniques requires optimum machine delivery performance. In this work, a monitoring system for multileaf collimator (MLC) performance, integrating semiautomated MLC quality control (QC) tests and statistical process control tools, was developed. The MLC performance monitoring system was used for almost a year on two commercially available MLC models. Control charts were used to establish MLC performance and assess test frequency required to achieve a given level of performance. MLC-related interlocks and servicing events were recorded during the monitoring period and were investigated as indicators of MLC performance variations. Methods: The QC test developed as part of the MLC performance monitoring system uses 2D megavoltage images (acquired using an electronic portal imaging device) of 23 fields to determine the location of the leaves with respect to the radiation isocenter. The precision of the MLC performance monitoring QC test and the MLC itself was assessed by detecting the MLC leaf positions on 127 megavoltage images of a static field. After initial calibration, the MLC performance monitoring QC test was performed 3–4 times/week over a period of 10–11 months to monitor positional accuracy of individual leaves for two different MLC models. Analysis of test results was performed using individuals control charts per leaf with control limits computed based on the measurements as well as two sets of specifications of ±0.5 and ±1 mm. Out-of-specification and out-of-control leaves were automatically flagged by the monitoring system and reviewed monthly by physicists. MLC-related interlocks reported by the linear accelerator and servicing events were recorded to help identify potential causes of nonrandom MLC leaf positioning variations. Results: The precision of the MLC performance monitoring QC test and the MLC itself was within ±0.22 mm for most MLC leaves

  10. Toward the development of intrafraction tumor deformation tracking using a dynamic multi-leaf collimator

    SciTech Connect

    Ge, Yuanyuan; O’Brien, Ricky T.; Shieh, Chun-Chien; Keall, Paul J.; Booth, Jeremy T.

    2014-06-15

    Purpose: Intrafraction deformation limits targeting accuracy in radiotherapy. Studies show tumor deformation of over 10 mm for both single tumor deformation and system deformation (due to differential motion between primary tumors and involved lymph nodes). Such deformation cannot be adapted to with current radiotherapy methods. The objective of this study was to develop and experimentally investigate the ability of a dynamic multi-leaf collimator (DMLC) tracking system to account for tumor deformation. Methods: To compensate for tumor deformation, the DMLC tracking strategy is to warp the planned beam aperture directly to conform to the new tumor shape based on real time tumor deformation input. Two deformable phantoms that correspond to a single tumor and a tumor system were developed. The planar deformations derived from the phantom images in beam's eye view were used to guide the aperture warping. An in-house deformable image registration software was developed to automatically trigger the registration once new target image was acquired and send the computed deformation to the DMLC tracking software. Because the registration speed is not fast enough to implement the experiment in real-time manner, the phantom deformation only proceeded to the next position until registration of the current deformation position was completed. The deformation tracking accuracy was evaluated by a geometric target coverage metric defined as the sum of the area incorrectly outside and inside the ideal aperture. The individual contributions from the deformable registration algorithm and the finite leaf width to the tracking uncertainty were analyzed. Clinical proof-of-principle experiment of deformation tracking using previously acquired MR images of a lung cancer patient was implemented to represent the MRI-Linac environment. Intensity-modulated radiation therapy (IMRT) treatment delivered with enabled deformation tracking was simulated and demonstrated. Results: The first

  11. A comparative study of the peripheral doses from a linear accelerator with a multileaf collimator system.

    PubMed

    Acun, Hediye; Zubaroglu, Ali; Kemikler, Gönül; Bozkurt, Ahmet

    2014-01-01

    This study presents a comparison of peripheral doses (PDs) measured using an ionisation chamber with treatment planning system (TPS) data and a Monte Carlo (MC) simulation of a 6-MV photon beam. The ion chamber measurements and MC simulation produced similar results for all out-of-field distances and field sizes considered in this study. For the 0° and 90° collimation angles, the average local per cent dose differences between the MC and TPS calculations were 2.7 % (range: -2.4, +22.6) and -1.7 % (range: -12.2, +10.8), respectively. The corresponding differences between the MC calculations and the ion chamber measurements were 2.2 % (range: -2.4, 24.7) and -1.8 % (range: -17, 15.2) for all field sizes and depths, respectively. Whereas the PDs increased with field sizes, the variations with depth were negligible at large distances. The TPS calculations usually yielded higher PDs than ion chamber measurements at distances close to the field edge. In contrast, at the farther distances, the TPS results indicated lower doses than both the ion chamber and the MC data. TPS data are not sufficient for use in calculating the out-of-field doses. These results can be used to estimate non-target organ doses to patients.

  12. NOTE: Adapting a generic BEAMnrc model of the BrainLAB m3 micro-multileaf collimator to simulate a local collimation device

    NASA Astrophysics Data System (ADS)

    Kairn, T.; Aland, T.; Franich, R. D.; Johnston, P. N.; Kakakhel, M. B.; Kenny, J.; Knight, R. T.; Langton, C. M.; Schlect, D.; Taylor, M. L.; Trapp, J. V.

    2010-09-01

    This work is focussed on developing a commissioning procedure so that a Monte Carlo model, which uses BEAMnrc's standard VARMLC component module, can be adapted to match a specific BrainLAB m3 micro-multileaf collimator (µMLC). A set of measurements are recommended, for use as a reference against which the model can be tested and optimized. These include radiochromic film measurements of dose from small and offset fields, as well as measurements of µMLC transmission and interleaf leakage. Simulations and measurements to obtain µMLC scatter factors are shown to be insensitive to relevant model parameters and are therefore not recommended, unless the output of the linear accelerator model is in doubt. Ultimately, this note provides detailed instructions for those intending to optimize a VARMLC model to match the dose delivered by their local BrainLAB m3 µMLC device.

  13. Neutron dosimetry in organs of an adult human phantom using linacs with multileaf collimator in radiotherapy treatments

    SciTech Connect

    Martinez-Ovalle, S. A.; Barquero, R.; Gomez-Ros, J. M.; Lallena, A. M.

    2012-05-15

    Purpose: To calculate absorbed doses due to neutrons in 87 organs/tissues for anthropomorphic phantoms, irradiated in position supine (head first into the gantry) with orientations anteroposterior (AP) and right-left (RLAT) with a 18 MV accelerator. Conversion factors from monitor units to {mu}Gy per neutron in organs, equivalent doses in organs/tissues, and effective doses, which permit to quantify stochastic risks, are estimated. Methods: MAX06 and FAX06 phantoms were modeled with MCNPX and irradiated with a 18 MV Varian Clinac 2100C/D accelerator whose geometry included a multileaf collimator. Two actual fields of a pelvic treatment were simulated using electron-photon-neutron coupled transport. Absorbed doses due to neutrons were estimated from kerma. Equivalent doses were estimated using the radiation weighting factor corresponding to an average incident neutron energy 0.47 MeV. Statistical uncertainties associated to absorbed doses, as calculated by MCNPX, were also obtained. Results: Largest doses were absorbed in shallowest (with respect to the neutron pathway) organs. In {mu}GyMU{sup -1}, values of 2.66 (for penis) and 2.33 (for testes) were found in MAX06, and 1.68 (for breasts), 1.05 (for lenses of eyes), and 0.94 (for sublingual salivary glands) in FAX06, in AP orientation. In RLAT, the largest doses were found for bone tissues (leg) just at the entrance of the beam in the body (right side in our case). Values, in {mu}GyMU{sup -1}, of 1.09 in upper leg bone right spongiosa, for MAX06, and 0.63 in mandible spongiosa, for FAX06, were found. Except for gonads, liver, and stomach wall, equivalent doses found for FAX06 were, in both orientations, higher than for MAX06. Equivalent doses in AP are higher than in RLAT for all organs/tissues other than brain and liver. Effective doses of 12.6 and 4.1 {mu}SvMU{sup -1} were found for AP and RLAT, respectively. The organs/tissues with larger relative contributions to the effective dose were testes and breasts, in

  14. Method for selecting minimum width of leaf in multileaf adjustable collimator while inhibiting passage of particle beams of radiation through sawtooth joints between collimator leaves

    DOEpatents

    Ludewigt, Bernhard; Bercovitz, John; Nyman, Mark; Chu, William

    1995-01-01

    A method is disclosed for selecting the minimum width of individual leaves of a multileaf adjustable collimator having sawtooth top and bottom surfaces between adjacent leaves of a first stack of leaves and sawtooth end edges which are capable of intermeshing with the corresponding sawtooth end edges of leaves in a second stack of leaves of the collimator. The minimum width of individual leaves in the collimator, each having a sawtooth configuration in the surface facing another leaf in the same stack and a sawtooth end edge, is selected to comprise the sum of the penetration depth or range of the particular type of radiation comprising the beam in the particular material used for forming the leaf; plus the total path length across all the air gaps in the area of the joint at the edges between two leaves defined between lines drawn across the peaks of adjacent sawtooth edges; plus at least one half of the length or period of a single sawtooth. To accomplish this, in accordance with the method of the invention, the penetration depth of the particular type of radiation in the particular material to be used for the collimator leaf is first measured. Then the distance or gap between adjoining or abutting leaves is selected, and the ratio of this distance to the height of the sawteeth is selected. Finally the number of air gaps through which the radiation will pass between sawteeth is determined by selecting the number of sawteeth to be formed in the joint. The measurement and/or selection of these parameters will permit one to determine the minimum width of the leaf which is required to prevent passage of the beam through the sawtooth joint.

  15. Dosimetric and radiobiological comparison of CyberKnife M6™ InCise multileaf collimator over IRIS™ variable collimator in prostate stereotactic body radiation therapy.

    PubMed

    Kathriarachchi, Vindu; Shang, Charles; Evans, Grant; Leventouri, Theodora; Kalantzis, Georgios

    2016-01-01

    The impetus behind our study was to establish a quantitative comparison between the IRIS collimator and the InCise multileaf collimator (MLC) (Accuray Inc. Synnyvale, CA) for prostate stereotactic body radiation therapy (SBRT). Treatment plans for ten prostate cancer patients were performed on MultiPlan™ 5.1.2 treatment planning system utilizing MLC and IRIS for 36.25 Gy in five fractions. To reduce the magnitude of variations between cases, the planning tumor volume (PTV) was defined and outlined for treating prostate gland only, assuming no seminal vesicle or ex-capsule involvement. Evaluation indices of each plan include PTV coverage, conformity index (CI), Paddick's new CI, homogeneity index, and gradient index. Organ at risk (OAR) dose sparing was analyzed by the bladder wall Dmax and V37Gy, rectum Dmax and V36Gy. The radiobiological response was evaluated by tumor control probability and normal tissue complication probability based on equivalent uniform dose. The dose delivery efficiency was evaluated on the basis of planned monitor units (MUs) and the reported treatment time per fraction. Statistical significance was tested using the Wilcoxon signed rank test. The studies indicated that CyberKnife M6™ IRIS and InCise™ MLC produce equivalent SBRT prostate treatment plans in terms of dosimetry, radiobiology, and OAR sparing, except that the MLC plans offer improvement of the dose fall-off gradient by 29% over IRIS. The main advantage of replacing the IRIS collimator with MLC is the improved efficiency, determined from the reduction of MUs by 42%, and a 36% faster delivery time. PMID:27217626

  16. Dosimetric and radiobiological comparison of CyberKnife M6™ InCise multileaf collimator over IRIS™ variable collimator in prostate stereotactic body radiation therapy

    PubMed Central

    Kathriarachchi, Vindu; Shang, Charles; Evans, Grant; Leventouri, Theodora; Kalantzis, Georgios

    2016-01-01

    The impetus behind our study was to establish a quantitative comparison between the IRIS collimator and the InCise multileaf collimator (MLC) (Accuray Inc. Synnyvale, CA) for prostate stereotactic body radiation therapy (SBRT). Treatment plans for ten prostate cancer patients were performed on MultiPlan™ 5.1.2 treatment planning system utilizing MLC and IRIS for 36.25 Gy in five fractions. To reduce the magnitude of variations between cases, the planning tumor volume (PTV) was defined and outlined for treating prostate gland only, assuming no seminal vesicle or ex-capsule involvement. Evaluation indices of each plan include PTV coverage, conformity index (CI), Paddick's new CI, homogeneity index, and gradient index. Organ at risk (OAR) dose sparing was analyzed by the bladder wall Dmax and V37Gy, rectum Dmax and V36Gy. The radiobiological response was evaluated by tumor control probability and normal tissue complication probability based on equivalent uniform dose. The dose delivery efficiency was evaluated on the basis of planned monitor units (MUs) and the reported treatment time per fraction. Statistical significance was tested using the Wilcoxon signed rank test. The studies indicated that CyberKnife M6™ IRIS and InCise™ MLC produce equivalent SBRT prostate treatment plans in terms of dosimetry, radiobiology, and OAR sparing, except that the MLC plans offer improvement of the dose fall-off gradient by 29% over IRIS. The main advantage of replacing the IRIS collimator with MLC is the improved efficiency, determined from the reduction of MUs by 42%, and a 36% faster delivery time. PMID:27217626

  17. Small field segments surrounded by large areas only shielded by a multileaf collimator: Comparison of experiments and dose calculation

    SciTech Connect

    Kron, T.; Clivio, A.; Vanetti, E.; Nicolini, G.; Cramb, J.; Lonski, P.; Cozzi, L.; Fogliata, A.

    2012-12-15

    Purpose: Complex radiotherapy fields delivered using a tertiary multileaf collimator (MLC) often feature small open segments surrounded by large areas of the beam only shielded by the MLC. The aim of this study was to test the ability of two modern dose calculation algorithms to accurately calculate the dose in these fields which would be common, for example, in volumetric modulated arc treatment (VMAT) and study the impact of variations in dosimetric leaf gap (DLG), focal spot size, and MLC transmission in the beam models. Methods: Nine test fields with small fields (0.6-3 cm side length) surrounded by large MLC shielded areas (secondary collimator 12 Multiplication-Sign 12 cm{sup 2}) were created using a 6 MV beam from a Varian Clinac iX linear accelerator with 120 leaf MLC. Measurements of output factors and profiles were performed using a diamond detector (PTW) and compared to two dose calculations algorithms anisotropic analytical algorithm [(AAA) and Acuros XB] implemented on a commercial radiotherapy treatment planning system (Varian Eclipse 10). Results: Both calculation algorithms predicted output factors within 1% for field sizes larger than 1 Multiplication-Sign 1 cm{sup 2}. For smaller fields AAA tended to underestimate the dose. Profiles were predicted well for all fields except for problems of Acuros XB to model the secondary penumbra between MLC shielded fields and the secondary collimator. A focal spot size of 1 mm or less, DLG 1.4 mm and MLC transmission of 1.4% provided a generally good model for our experimental setup. Conclusions: AAA and Acuros XB were found to predict the dose under small MLC defined field segments well. While DLG and focal spot affect mostly the penumbra, the choice of correct MLC transmission will be essential to model treatments such as VMAT accurately.

  18. Multileaf Collimator Tracking Improves Dose Delivery for Prostate Cancer Radiation Therapy: Results of the First Clinical Trial

    SciTech Connect

    Colvill, Emma; Booth, Jeremy T.; O'Brien, Ricky T.; Eade, Thomas N.; Kneebone, Andrew B.; Poulsen, Per R.; Keall, Paul J.

    2015-08-01

    Purpose: To test the hypothesis that multileaf collimator (MLC) tracking improves the consistency between the planned and delivered dose compared with the dose without MLC tracking, in the setting of a prostate cancer volumetric modulated arc therapy trial. Methods and Materials: Multileaf collimator tracking was implemented for 15 patients in a prostate cancer radiation therapy trial; in total, 513 treatment fractions were delivered. During each treatment fraction, the prostate trajectory and treatment MLC positions were collected. These data were used as input for dose reconstruction (multiple isocenter shift method) to calculate the treated dose (with MLC tracking) and the dose that would have been delivered had MLC tracking not been applied (without MLC tracking). The percentage difference from planned for target and normal tissue dose-volume points were calculated. The hypothesis was tested for each dose-volume value via analysis of variance using the F test. Results: Of the 513 fractions delivered, 475 (93%) were suitable for analysis. The mean difference and standard deviation between the planned and treated MLC tracking doses and the planned and without-MLC tracking doses for all 475 fractions were, respectively, PTV D{sub 99%} −0.8% ± 1.1% versus −2.1% ± 2.7%; CTV D{sub 99%} −0.6% ± 0.8% versus −0.6% ± 1.1%; rectum V{sub 65%} 1.6% ± 7.9% versus −1.2% ± 18%; and bladder V{sub 65%} 0.5% ± 4.4% versus −0.0% ± 9.2% (P<.001 for all dose-volume results). Conclusion: This study shows that MLC tracking improves the consistency between the planned and delivered doses compared with the modeled doses without MLC tracking. The implications of this finding are potentially improved patient outcomes, as well as more reliable dose-volume data for radiobiological parameter determination.

  19. An alternative effective method for verifying the multileaf collimator leaves speed by using a digital-video imaging system

    NASA Astrophysics Data System (ADS)

    Hwang, Ing-Ming; Wu, Jay; Chuang, Keh-Shih; Ding, Hueisch-Jy

    2010-11-01

    We present an alternative effective method for verifying the multileaf collimator (MLC) leaves speed using a digital-video imaging system in daily dynamic conformal radiation therapy (DCRT) and intensity-modulation radiation therapy (IMRT) in achieving increased convenience and shorter treatment times. The horizontal leaves speed measured was within 1.76-2.08 cm/s. The mean full range of traveling time was 20 s. The initial speed-up time was within 1.5-2.0 s, and the slowing-down time was within 2.0-2.5 s. Due to gravity the maximum speed-up effect in the X1 bank was +0.10 cm/s, but the lagging effect in the X2 bank was -0.20 cm/s. This technique offered an alternative method with electronic portal imaging device (EPID), charged coupled device (CCD) or a light field for the measurement of MLC leaves speed. When time taken on the linac was kept to a minimum, the image could be processed off-line.

  20. Rounded leaf end effect of multileaf collimator on penumbra width and radiation field offset: an analytical and numerical study

    PubMed Central

    Zhou, Dong; Zhang, Hui; Ye, Peiqing

    2015-01-01

    Background Penumbra characteristics play a significant role in dose delivery accuracy for radiation therapy. For treatment planning, penumbra width and radiation field offset strongly influence target dose conformity and organ at risk sparing. Methods In this study, we present an analytical and numerical approach for evaluation of the rounded leaf end effect on penumbra characteristics. Based on the rule of half-value layer, algorithms for leaf position calculation and radiation field offset correction were developed, which were advantageous particularly in dealing with large radius leaf end. Computer simulation was performed based on the Monte Carlo codes of EGSnrc/BEAMnrc, with groups of leaf end radii and source sizes. Data processing technique of curve fitting was employed for deriving penumbra width and radiation field offset. Results Results showed that penumbra width increased with source size. Penumbra width curves for large radius leaf end were U-shaped. This observation was probably related to the fact that radiation beams penetrated through the proximal and distal leaf sides. In contrast, source size had negligible impact on radiation field offset. Radiation field offsets were found to be constant both for analytical method and numerical simulation. However, the overall resulting values of radiation field offset obtained by analytical method were slightly smaller compared with Monte Carlo simulation. Conclusions The method we proposed could provide insight into the investigation of rounded leaf end effects on penumbra characteristics. Penumbra width and radiation field offset calibration should be carefully performed to commission multileaf collimator for intensity modulated radiotherapy. PMID:26401137

  1. Lateral loss and dose discrepancies of multileaf collimator segments in intensity modulated radiation therapy.

    PubMed

    Cheng, Chee W; Das, Indra J; Huq, M Saiful

    2003-11-01

    In the step-and-shoot technique delivery of intensity modulated radiation therapy (IMRT), each static field consists of a number of beamlets, some of which may be very small. In this study, we measured the dose characteristics for a range of field sizes: 2 x 2 to 12 x 10 cm2 for 6 and 15 MV x rays. For a given field length, a number of treatment fields are set up by sequentially increasing the field width using a multi leaf collimator. A set of fields is delivered with the accelerator operated in the IMRT mode. Using an ion chamber, the output factors at 1 cm and 3 cm laterally from a field edge are measured at different depths in a solid water phantom. Our results show that with insufficient lateral distance in at least one direction, the absorbed dose never reaches the equilibrium values, and can be significantly lower for very small field sizes. For example, the output factor of the 2 x 2 cm2 field relative to 10 x 10 cm2 at d(max0 is 0.832 and 0.790 for 6 MV and 15 MV x rays, respectively. Multiple output factor curves are obtained for different field lengths and different buildup conditions. Thus under nonequilibrium conditions, output factors are critically dependent on the field size and the conventional method of determining the equivalent square does not apply. Comparison of output factors acquired in the commissioning of the accelerator with those measured in the present study under conditions of nonequilibrium shows large discrepancies between the two sets of measurements. Thus monitor units generated by a treatment planning system using beam data commissioned with symmetric fields may be underestimated by > 5%, depending on the size and shape of the segments. To facilitate manual MU calculation as an independent check in step-and-shoot IMRT, the concept of effective equivalent square (EES) is introduced. Using EES, output factors can be calculated using existing beam data for fields with asymmetric collimator settings and under conditions of lateral

  2. Development and commissioning of a multileaf collimator model in Monte Carlo dose calculations for intensity-modulated radiation therapy

    SciTech Connect

    Jang Siyoung; Vassiliev, Oleg N.; Liu, H. Helen; Mohan, Radhe; Siebers, Jeffrey V.

    2006-03-15

    A multileaf collimator (MLC) model, 'MATMLC', was developed to simulate MLCs for Monte Carlo (MC) dose calculations of intensity-modulated radiation therapy (IMRT). This model describes MLCs using matrices of regions, each of which can be independently defined for its material and geometry, allowing flexibility in simulating MLCs from various manufacturers. The free parameters relevant to the dose calculations with this MLC model included MLC leaf density, interleaf air gap, and leaf geometry. To commission the MLC model and its free parameters for the Varian Millennium MLC-120 (Varian Oncology Systems, Palo Alto, CA), we used the following leaf patterns: (1) MLC-blocked fields to test the effects of leaf transmission and leakage; (2) picket-fence fields to test the effects of the interleaf air gap and tongue-groove design; and (3) abutting-gap fields to test the effects of rounded leaf ends. Transmission ratios and intensity maps for these leaf patterns were calculated with various sets of modeling parameters to determine their dosimetric effects, sensitivities, and their optimal combinations to give the closest agreement with measured results. Upon commissioning the MLC model, we computed dose distributions for clinical IMRT plans using the MC system and verified the results with those from ion chamber and thermoluminescent dosimeter measurements in water phantoms and anthropomorphic phantoms. This study showed that the MLC transmission ratios were strongly dependent on both leaf density and the interleaf air gap. The effect of interleaf air gap and tongue-groove geometry can be determined most effectively through fence-type MLC patterns. Using the commissioned MLC model, we found that the calculated dose from the MC system agreed with the measured data within clinically acceptable criteria from low- to high-dose regions, showing that the model is acceptable for clinical applications.

  3. SU-E-I-49: The Evaluation of Usability of Multileaf Collimator for Diagnostic Radiation in Cephalometric Exposure

    SciTech Connect

    Han, S; Kim, K; Jung, H; Kim, M; Ji, Y; Park, S; Choi, S

    2014-06-01

    Purpose: This study evaluated usability of Multileaf collimator (MLC) for diagnostic radiation in cephalometric exposure using optical stimulated luminance dosimeters (OSLDs) Methods: The MLC material was made alloy tool steel (SKD-11) and the density of it is 7.89g/m3 that is similar to it of steel (Fe, 7.85 g/m3) and the MLC was attached to general radiography unit (Rex-650R, Listem Inc, Korea) for cephalometric exposure. The OSLDs that used were nanoDotTM Dosimeter (Landauer Inc, Glenwood, USA) and we read out OSLDs with micro star system (Landauer Inc, Glenwood, USA). The Optical annealing system contained fluorescent lamps (Osram lumilux, 24 W, 280 ∼780 nm). To measure absorbed dose using OSLDs, was carried out dosimetric characteristics of OSLDs. Based on these, we evaluated dose reduction of critical organ (Eyes, Thyroids) with MLC in cephalometric exposure Results: The dosimetric characteristics were following that batch homogeneity was 1.21% and reproducibility was 0.96% of the coefficient of variation The linearity was that the correlation of between dose and count was fitted by linear function (dose,mGy = 0.00029 × Count, R2 =0.997). The range of angular dependence was from −3.6% to 3.7% variation when each degree was normalized by zero degree. The organ dose of Rt. eye, Lt eye, thyroids were 77.8 μGy, 337.0 μGy, 323.1μGy, respectively in open field and the dose reduction of organ dose was 10.6%(8.3μGy), 12.4 %(42 μGy), 87.1%(281.4μGy) with MLC Conclusion: We certified dose reduction of organ dose in cephalometric exposure. The dose reduction of Eye was 11% because of reduction of field size and it of thyroids was 87% by primary beam shielding.

  4. Development, physical properties and clinical applicability of a mechanical Multileaf Collimator for the use in Cobalt-60 radiotherapy.

    PubMed

    Langhans, Marco; Echner, Gernot; Runz, Armin; Baumann, Martin; Xu, Mark; Ueltzhöffer, Stefan; Häring, Peter; Schlegel, Wolfgang

    2015-04-21

    According to the Directory of Radiotherapy Centres (DIRAC) there are 2348 Cobalt-60 (Co-60) teletherapy units worldwide, most of them in low and middle income countries, compared to 11046 clinical accelerators. To improve teletherapy with Co-60, a mechanical Multi-Leaf Collimator (MLC) was developed, working with pneumatic pressure and thus independent of electricity supply. Instead of tungsten, brass was used as leaf material to make the mechanical MLC more affordable. The physical properties and clinical applicability of this mechanical MLC are presented here. The leakage strongly depends on the fieldsize of the therapy unit due to scatter effects. The maximum transmission through the leaves measured 2.5 cm from the end-to-end gap, within a field size of 20 cm × 30 cm defined by jaws of the therapy unit at 80 cm SAD, amounts 4.2%, normalized to an open 10 cm × 10 cm field, created by the mechanical MLC. Within a precollimated field size of 12.5 cm × 12.5 cm, the end-to-end leakage is 6.5% normalized to an open 10 cm × 10 cm field as well. This characteristic is clinically acceptable considering the criteria for non-IMRT MLCs of the International Electrotechnical Commission (IEC 60601-2-1). The penumbra for a 10 cm × 10 cm field was measured to be 9.14 mm in plane and 8.38 mm cross plane. The clinical applicability of the designed mechanical MLC was affirmed by measurements relating to all relevant clinical properties such as penumbra, leakage, output factors and field widths. Hence this novel device presents an apt way forward to make radiotherapy with conformal fields possible in low-infrastructure environments, using gantry based Co-60 therapy units.

  5. Determining the optimal dosimetric leaf gap setting for rounded leaf-end multileaf collimator systems by simple test fields.

    PubMed

    Yao, Weiguang; Farr, Jonathan B

    2015-01-01

    Individual QA for IMRT/VMAT plans is required by protocols. Sometimes plans cannot pass the institute's QA criteria. For the Eclipse treatment planning system (TPS) with rounded leaf-end multileaf collimator (MLC), one practical way to improve the agreement of planned and delivered doses is to tune the value of dosimetric leaf gap (DLG) in the TPS from the measured DLG. We propose that this step may be necessary due to the complexity of the MLC system, including dosimetry of small fields and the tongue-and-groove (T&G) effects, and report our use of test fields to obtain linac-specific optimal DLGs in TPSs. More than 20 original patient plans were reoptimized with the linac-specific optimal DLG value. We examined the distribution of gaps and T&G extensions in typical patient plans and the effect of using the optimal DLG on the distribution. The QA pass rate of patient plans using the optimal DLG was investigated. The dose-volume histograms (DVHs) of targets and organs at risk were checked. We tested three MLC systems (Varian millennium 120 MLC, high-definition 120 MLC, and Siemens 160 MLC) installed in four Varian linear accelerators (linacs) (TrueBEAM STx, Trilogy, Clinac 2300 iX, and Clinac 21 EX) and 1 Siemens linac (Artiste). With an optimal DLG, the individual QA for all those patient plans passed the institute's criteria (95% in DTA test or gamma test with 3%/3 mm/10%), even though most of these plans had failed to pass QA when using original DLGs optimized from typical patient plans or from the optimization process (automodeler) of Pinnacle TPS. Using either our optimal DLG or one optimized from typical patient plans or from the Pinnacle optimization process yielded similar DVHs. PMID:26218999

  6. SU-E-T-214: Intensity Modulated Proton Therapy (IMPT) Based On Passively Scattered Protons and Multi-Leaf Collimation: Prototype TPS and Dosimetry Study

    SciTech Connect

    Sanchez-Parcerisa, D; Carabe-Fernandez, A

    2014-06-01

    Purpose. Intensity-modulated proton therapy is usually implemented with multi-field optimization of pencil-beam scanning (PBS) proton fields. However, at the view of the experience with photon-IMRT, proton facilities equipped with double-scattering (DS) delivery and multi-leaf collimation (MLC) could produce highly conformal dose distributions (and possibly eliminate the need for patient-specific compensators) with a clever use of their MLC field shaping, provided that an optimal inverse TPS is developed. Methods. A prototype TPS was developed in MATLAB. The dose calculation process was based on a fluence-dose algorithm on an adaptive divergent grid. A database of dose kernels was precalculated in order to allow for fast variations of the field range and modulation during optimization. The inverse planning process was based on the adaptive simulated annealing approach, with direct aperture optimization of the MLC leaves. A dosimetry study was performed on a phantom formed by three concentrical semicylinders separated by 5 mm, of which the inner-most and outer-most were regarded as organs at risk (OARs), and the middle one as the PTV. We chose a concave target (which is not treatable with conventional DS fields) to show the potential of our technique. The optimizer was configured to minimize the mean dose to the OARs while keeping a good coverage of the target. Results. The plan produced by the prototype TPS achieved a conformity index of 1.34, with the mean doses to the OARs below 78% of the prescribed dose. This Result is hardly achievable with traditional conformal DS technique with compensators, and it compares to what can be obtained with PBS. Conclusion. It is certainly feasible to produce IMPT fields with MLC passive scattering fields. With a fully developed treatment planning system, the produced plans can be superior to traditional DS plans in terms of plan conformity and dose to organs at risk.

  7. SU-E-T-610: Comparison of Treatment Times Between the MLCi and Agility Multileaf Collimators

    SciTech Connect

    Ramsey, C; Bowling, J

    2014-06-01

    Purpose: The Agility is a new 160-leaf MLC developed by Elekta for use in their Infinity and Versa HD linacs. As compared to the MLCi, the Agility increased the maximum leaf speed from 2 cm/s to 3.5 cm/s, and the maximum primary collimator speed from 1.5 cm/s to 9.0 cm/s. The purpose of this study was to determine if the Agility MLC resulted in improved plan quality and/or shorter treatment times. Methods: An Elekta Infinity that was originally equipped with a 80 leaf MLCi was upgraded to an 160 leaf Agility. Treatment plan quality was evaluated using the Pinnacle planning system with SmartArc. Optimization was performed once for the MLCi and once for the Agility beam models using the same optimization parameters and the same number of iterations. Patient treatment times were measured for all IMRT, VMAT, and SBRT patients treated on the Infinity with the MLCi and Agility MLCs. Treatment times were extracted from the EMR and measured from when the patient first walked into the treatment room until exiting the treatment room. Results: 11,380 delivery times were measured for patients treated with the MLCi, and 1,827 measurements have been made for the Agility MLC. The average treatment times were 19.1 minutes for the MLCi and 20.8 minutes for the Agility. Using a t-test analysis, there was no difference between the two groups (t = 0.22). The dose differences between patients planned with the MLCi and the Agility MLC were minimal. For example, the dose difference for the PTV, GTV, and cord for a head and neck patient planned using Pinnacle were effectively equivalent. However, the dose to the parotid glands was slightly worse with the Agility MLC. Conclusion: There was no statistical difference in treatment time, or any significant dosimetric difference between the Agility MLC and the MLCi.

  8. Real-Time Target Position Estimation Using Stereoscopic Kilovoltage/Megavoltage Imaging and External Respiratory Monitoring for Dynamic Multileaf Collimator Tracking

    SciTech Connect

    Cho, Byungchul; Poulsen, Per Rugaard; Sawant, Amit; Ruan, Dan; Keall, Paul J.

    2011-01-01

    Purpose: To develop a real-time target position estimation method using stereoscopic kilovoltage (kV)/megavoltage (MV) imaging and external respiratory monitoring, and to investigate the performance of a dynamic multileaf collimator tracking system using this method. Methods and Materials: The real-time three-dimensional internal target position estimation was established by creating a time-varying correlation model that connected the external respiratory signals with the internal target motion measured intermittently using kV/MV imaging. The method was integrated into a dynamic multileaf collimator tracking system. Tracking experiments were performed for 10 thoracic/abdominal traces. A three-dimensional motion platform carrying a gold marker and a separate one-dimensional motion platform were used to reproduce the target and external respiratory motion, respectively. The target positions were detected by kV (1 Hz) and MV (5.2 Hz) imaging, and external respiratory motion was captured by an optical system (30 Hz). The beam-target alignment error was quantified as the positional difference between the target and circular beam center on the MV images acquired during tracking. The correlation model error was quantified by comparing a model estimate and measured target positions. Results: The root-mean-square errors in the beam-target alignment that had ranged from 3.1 to 7.6 mm without tracking were reduced to <1.5 mm with tracking, except during the model building period (6 s). The root-mean-square error in the correlation model was submillimeters in all directions. Conclusion: A novel real-time target position estimation method was developed and integrated into a dynamic multileaf collimator tracking system and demonstrated an average submillimeter geometric accuracy after initializing the internal/external correlation model. The method used hardware tools available on linear accelerators and therefore shows promise for clinical implementation.

  9. Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT)

    PubMed Central

    Wang, Iris Z.; Kumaraswamy, Lalith K.; Podgorsak, Matthew B.

    2016-01-01

    Background This study is to report 1) the sensitivity of intensity modulated radiation therapy (IMRT) QA method for clinical volumetric modulated arc therapy (VMAT) plans with multi-leaf collimator (MLC) leaf errors that will not trigger MLC interlock during beam delivery; 2) the effect of non-beam-hold MLC leaf errors on the quality of VMAT plan dose delivery. Materials and methods. Eleven VMAT plans were selected and modified using an in-house developed software. For each control point of a VMAT arc, MLC leaves with the highest speed (1.87-1.95 cm/s) were set to move at the maximal allowable speed (2.3 cm/s), which resulted in a leaf position difference of less than 2 mm. The modified plans were considered as ‘standard’ plans, and the original plans were treated as the ‘slowing MLC’ plans for simulating ‘standard’ plans with leaves moving at relatively lower speed. The measurement of each ‘slowing MLC’ plan using MapCHECK®2 was compared with calculated planar dose of the ‘standard’ plan with respect to absolute dose Van Dyk distance-to-agreement (DTA) comparisons using 3%/3 mm and 2%/2 mm criteria. Results All ‘slowing MLC’ plans passed the 90% pass rate threshold using 3%/3 mm criteria while one brain and three anal VMAT cases were below 90% with 2%/2 mm criteria. For ten out of eleven cases, DVH comparisons between ‘standard’ and ‘slowing MLC’ plans demonstrated minimal dosimetric changes in targets and organs-at-risk. Conclusions For highly modulated VMAT plans, pass rate threshold (90%) using 3%/3mm criteria is not sensitive in detecting MLC leaf errors that will not trigger the MLC leaf interlock. However, the consequential effects of non-beam hold MLC errors on target and OAR doses are negligible, which supports the reliability of current patient-specific IMRT quality assurance (QA) method for VMAT plans. PMID:27069458

  10. Dosimetric comparison of different multileaf collimator leaves in treatment planning of intensity-modulated radiotherapy for cervical cancer

    SciTech Connect

    Wang, Shichao; Ai, Ping; Xie, Li; Xu, Qingfeng; Bai, Sen; Lu, You; Li, Ping; Chen, Nianyong

    2013-01-01

    To study the effect of multileaf collimator (MLC) leaf widths (standard MLC [sMLC] width of 10 mm and micro-MLC [mMLC] width of 4 mm) on intensity-modulated radiotherapy (IMRT) for cervical cancer. Between January 2010 and August 2010, a retrospective analysis was conducted on 12 patients with cervical cancer. The treatment plans for all patients were generated with the same machine setup parameters and optimization methods in a treatment planning system (TPS) based on 2 commercial Elekta MLC devices. The dose distribution for the planning tumor volume (PTV), the dose sparing for organs at risk (OARs), the monitor units (MUs), and the number of IMRT segments were evaluated. For the delivery efficiency, the MUs were significantly higher in the sMLC-IMRT plan than in the mMLC-IMRT plan (802 ± 56.9 vs 702 ± 56.7; p < 0.05). The number of segments in the plans were 58.75 ± 1.8 and 59 ± 1.04 (p > 0.05). For the planning quality, the conformity index (CI) between the 2 paired IMRT plans with the mMLC and the sMLC did not differ significantly (average: 0.817 ± 0.024 vs 0.810 ± 0.028; p > 0.05). The differences of the homogeneity index (HI) between the 2 paired plans were statistically significant (average: 1.122 ± 0.010 vs 1.132 ± 0.014; p < 0.01). For OARs, the rectum, bladder, small intestine, and bony pelvis were evaluated in terms of V{sub 10}, V{sub 20}, V{sub 30}, and V{sub 40}, percentage of contoured OAR volumes receiving 10, 20, 30, and 40 Gy, respectively, and the mean dose (D{sub mean}) received. The IMRT plans with the mMLC protected the OARs better than the plans with the sMLC. There were significant differences (p < 0.05) in evaluated parameters between the 2 paired IMRT plans, except for V{sub 30} and V{sub 40} of the rectum and V{sub 10}, V{sub 20}, V{sub 40}, and D{sub mean} of the bladder. IMRT plans with the mMLC showed advantages over the plans with the sMLC in dose homogeneity for targets, dose sparing of OARs, and fewer MUs in cervical cancer.

  11. SU-E-T-348: Verification MU Calculation for Conformal Radiotherapy with Multileaf Collimator Using Report AAPM TG 114

    SciTech Connect

    Adrada, A; Tello, Z; Medina, L; Garrigo, E; Venencia, D

    2014-06-01

    Purpose: The purpose of this work was to develop and validate an open source independent MU dose calculation software for 3D conformal radiotherapy with multileaf high and low resolution according to the report of AAPM TG 11 Methods: Treatment plans were done using Iplan v4.5 BrainLAB TPS. A 6MV photon beam produced by Primus and Novalis linear accelerators equipped with an Optifocus MLC and HDMLC, respectively. TPS dose calculation algorithms were pencil beam and Monte Carlo. 1082 treatments plans were selected for the study. The algorithm was written in free and open source CodeBlocks C++ platform. Treatment plans were imported by the software using RTP format. Equivalent size field is obtained from the positions of the leaves; the effective depth of calculation can be introduced by TPS's dosimetry report or automatically calculated starting from SSD. The inverse square law is calculated by the 3D coordinates of the isocenter and normalization point of the treatment plan. The dosimetric parameters TPR, Sc, Sp and WF are linearly interpolated. Results: 1082 plans of both machines were analyzed. The average uncertainty between the TPS and the independent calculation was −0.43% ± 2.42% [−7.90%, 7.50%]. Specifically for the Primus the variation obtained was −0.85% ± 2.53% and for the Novalis 0.00% ± 2.23%. Data show that 94.8% of the cases the uncertainty was less than or equal to 5%, while 98.9% is less than or equal to 6%. Conclusion: The developed software is appropriate for use in calculation of UM. This software can be obtained upon request.

  12. Photonic crystals for broadband, omnidirectional self-collimation

    NASA Astrophysics Data System (ADS)

    Chuang, Y.-C.; Suleski, T. J.

    2011-03-01

    In this paper, multiple photonic crystal (PC) structures are proposed to improve three-dimensional (3D) self-collimation performance, including two 3D PCs (tetragonal lattice structures and a complex hexagonal lattice structure) and two two-dimensional (2D) PCs (triangular lattice structures and kagome lattice structures) with out-of-plane orientation. Different design strategies are investigated and compared in terms of the resulting self-collimation performance. Several desired 3D properties are numerically realized for the first time, including broadband 3D self-collimation, omnidirectional beam confinement and broadband omnidirectional self-collimation. These developments can enable future self-collimation applications, such as multiplexers, PC core fibers and solar light collection.

  13. Comparison of Monte Carlo collimator transport methods for photon treatment planning in radiotherapy

    SciTech Connect

    Schmidhalter, D.; Manser, P.; Frei, D.; Volken, W.; Fix, M. K.

    2010-02-15

    Purpose: The aim of this work was a Monte Carlo (MC) based investigation of the impact of different radiation transport methods in collimators of a linear accelerator on photon beam characteristics, dose distributions, and efficiency. Thereby it is investigated if it is possible to use different simplifications in the radiation transport for some clinical situations in order to save calculation time. Methods: Within the Swiss Monte Carlo Plan, a GUI-based framework for photon MC treatment planning, different MC methods are available for the radiation transport through the collimators [secondary jaws and multileaf collimator (MLC)]: EGSnrc (reference), VMC++, and Pin (an in-house developed MC code). Additional nonfull transport methods were implemented in order to provide different complexity levels for the MC simulation: Considering collimator attenuation only, considering Compton scatter only or just the firstCompton process, and considering the collimators as totally absorbing. Furthermore, either a simple or an exact geometry of the collimators can be selected for the absorbing or attenuation method. Phasespaces directly above and dose distributions in a water phantom are analyzed for academic and clinical treatment fields using 6 and 15 MV beams, including intensity modulated radiation therapy with dynamic MLC. Results: For all MC transport methods, differences in the radial mean energy and radial energy fluence are within 1% inside the geometric field. Below the collimators, the energy fluence is underestimated for nonfull MC transport methods ranging from 5% for Compton to 100% for Absorbing. Gamma analysis using EGSnrc calculated doses as reference shows that the percentage of voxels fulfilling a 1% /1 mm criterion is at least 98% when using VMC++, Compton, or firstCompton transport methods. When using the methods Pin, Transmission, Flat-Transmission, Flat-Absorbing or Absorbing, the mean value of points fulfilling this criterion over all tested cases is 97

  14. Dosimetric characterization of a multileaf collimator for a new four-dimensional image-guided radiotherapy system with a gimbaled x-ray head, MHI-TM2000

    SciTech Connect

    Nakamura, Mitsuhiro; Sawada, Akira; Ishihara, Yoshitomo; Takayama, Kenji; Mizowaki, Takashi; Kaneko, Shuji; Yamashita, Mikiko; Tanabe, Hiroaki; Kokubo, Masaki; Hiraoka, Masahiro

    2010-09-15

    Purpose: To present the dosimetric characterization of a multileaf collimator (MLC) for a new four-dimensional image-guided radiotherapy system with a gimbaled x-ray head, MHI-TM2000. Methods: MHI-TM2000 has an x-ray head composed of an ultrasmall linear accelerator guide and a system-specific MLC. The x-ray head can rotate along the two orthogonal gimbals (pan and tilt rotations) up to {+-}2.5 deg., which swings the beam up to {+-}41.9 mm in each direction from the isocenter on the isocenter plane perpendicular to the beam. The MLC design is a single-focus type, has 30 pairs of 5 mm thick leaves at the isocenter, and produces a maximum field size of 150x150 mm{sup 2}. Leaf height and length are 110 and 260 mm, respectively. Each leaf end is circular, with a radius of curvature of 370 mm. The distance that each leaf passes over the isocenter is 77.5 mm. Radiation leakage between adjacent leaves is minimized by an interlocking tongue-and-groove (T and G) arrangement with the height of the groove part 55 mm. The dosimetric characterizations including field characteristics, leaf position accuracy, leakage, and T and G effect were evaluated using a well-commissioned 6 MV photon beam, EDR2 films (Kodak, Rochester, NY), and water-equivalent phantoms. Furthermore, the field characteristics and leaf position accuracy were evaluated under conditions of pan or tilt rotation. Results: The differences between nominal and measured field sizes were within {+-}0.5 mm. Although the penumbra widths were greater with wider field size, the maximum width was <5.5 mm even for the fully opened field. Compared to the results of field characteristics without pan or tilt rotation, the variation in field size, penumbra width, flatness, and symmetry was within {+-}1 mm/1% at the maximum pan or tilt rotational angle. The leaf position accuracy was 0.0{+-}0.1 mm, ranging from -0.3 to 0.2 mm at four gantry angles of 0 deg., 90 deg., 180 deg., and 270 deg. with and without pan or tilt rotation

  15. Dosimetric effects of multileaf collimator leaf width on intensity-modulated radiotherapy for head and neck cancer

    SciTech Connect

    Hong, Chae-Seon; Ju, Sang Gyu Kim, Minkyu; Kim, Jin Man; Han, Youngyih; Ahn, Yong Chan; Choi, Doo Ho; Park, Hee Chul; Kim, Jung-in; Nam, Heerim; Suh, Tae-Suk

    2014-02-15

    Purpose: The authors evaluated the effects of multileaf collimator (MLC) leaf width (2.5 vs. 5 mm) on dosimetric parameters and delivery efficiencies of intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) for head and neck (H and N) cancers. Methods: The authors employed two types of mock phantoms: large-sized head and neck (LH and N) and small-sized C-shape (C-shape) phantoms. Step-and-shoot IMRT (S and S-IMRT) and VMAT treatment plans were designed with 2.5- and 5.0-mm MLC for both C-shape and LH and N phantoms. Their dosimetric characteristics were compared in terms of the conformity index (CI) and homogeneity index (HI) for the planning target volume (PTV), the dose to organs at risk (OARs), and the dose-spillage volume. To analyze the effects of the field and arc numbers, 9-field IMRT (9F-IMRT) and 13-field IMRT (13F-IMRT) plans were established for S and S-IMRT. For VMAT, single arc (VMAT{sub 1}) and double arc (VMAT{sub 2}) plans were established. For all plans, dosimetric verification was performed using the phantom to examine the relationship between dosimetric errors and the two leaf widths. Delivery efficiency of the two MLCs was compared in terms of beam delivery times, monitor units (MUs) per fraction, and the number of segments for each plan. Results: 2.5-mm MLC showed better dosimetric characteristics in S and S-IMRT and VMAT for C-shape, providing better CI for PTV and lower spinal cord dose and high and intermediate dose-spillage volume as compared with the 5-mm MLC (p < 0.05). However, no significant dosimetric benefits were provided by the 2.5-mm MLC for LH and N (p > 0.05). Further, beam delivery efficiency was not observed to be significantly associated with leaf width for either C-shape or LH and N. However, MUs per fraction were significantly reduced for the 2.5-mm MLC for the LH and N. In dosimetric error analysis, absolute dose evaluations had errors of less than 3%, while the Gamma passing rate was

  16. Photonic crystal composites-based wide-band optical collimator

    NASA Astrophysics Data System (ADS)

    Shi, Jinjie; Juluri, Bala Krishna; Lin, Sz-Chin Steven; Lu, Mengqian; Gao, Tieyu; Huang, Tony Jun

    2010-08-01

    Photonic crystal (PC) composites are sequenced series of PCs that feature the same periods but different filling fractions. By properly tuning the filling fractions of the individual PCs and merging the working band of each PC into a continuous frequency range, wide-band self-collimation of optical signals can be realized. The band diagrams and the equal-frequency contours of the PC structures were calculated through the plane wave expansion method and the finite-difference time-domain method was employed to simulate the propagation of electromagnetic waves through the PC structures. Our results show that while a single PC can only collimate optical waves over a narrow frequency range, a PC composite exhibits a much wider collimation band. Such a wide-band optical collimation lens can be useful in applications that demand directional optical energy flow over a long distance, such as optical imaging and biosensing.

  17. Minimum requirements for commissioning and long-term quality assurance of Elekta multi-leaf collimator for volumetric modulated arc therapy.

    PubMed

    Tatsumi, Daisaku; Nakada, Ryosei; Yomoda, Akane; Ishii, Kentaro; Tsutsumi, Shinichi; Inoue, Makoto; Ichida, Takao; Hosono, Masako N; Miki, Yukio

    2013-01-01

    We have proposed minimum requirements for commissioning and long-term quality assurance (QA) of an Elekta multi-leaf collimator (MLC) for volumetric modulated arc therapy (VMAT). The MLC leaf position accuracy during VMAT delivery was evaluated with the use of three different QA test plans: (1) a leaf gap-width test between opposing leaves by measurement of the isocenter dose during constant-gap sliding-window delivery with varied dose rates, MLC leaf speeds, and gantry angles; (2) a leaf position test by picket-fence delivery with and without gantry rotation; and (3) a leaf-bank symmetry test by measurement of the field geometry with different collimator angles at a fixed gantry position. All the QA test plans were created using an ERGO++ treatment-planning system. The leaf gap-width deviation was within 0.2 mm, the leaf position deviation was within 0.5 mm, and the leaf-bank symmetry error was within 0.5 mm under all the test conditions. MLC leaf position accuracy and long-term stability were confirmed by the proposed procedures.

  18. SU-E-T-515: Field-In-Field Compensation Technique Using Multi-Leaf Collimator to Deliver Total Body Irradiation (TBI) Dose

    SciTech Connect

    Lakeman, T; Wang, IZ

    2014-06-01

    Purpose: Total body irradiation (TBI) uses large parallel-opposed radiation fields to suppress the patient's immune system and eradicate the residual cancer cells in preparation of recipient for bone marrow transplant. The manual placement of lead compensators has been used conventionally to compensate for the varying thickness through the entire body in large-field TBI. The goal of this study is to pursue utilizing the modern field-in-field (FIF) technique with the multi-leaf collimator (MLC) to more accurately and efficiently deliver dose to patients in need of TBI. Method: Treatment plans utilizing the FIF technique to deliver a total body dose were created retrospectively for patients for whom CT data had been previously acquired. Treatment fields include one pair of opposed open large fields (collimator=45°) with a specific weighting and a succession of smaller fields (collimator=90°) each with their own weighting. The smaller fields are shaped by moving MLC to block the sections of the patient which have already received close to 100% of the prescribed dose. The weighting factors for each of these fields were calculated using the attenuation coefficient of the initial lead compensators and the separation of the patient in different positions in the axial plane. Results: Dose-volume histograms (DVH) were calculated for evaluating the FIF compensation technique. The maximum body doses calculated from the DVH were reduced from the non-compensated 179.3% to 148.2% in the FIF plans, indicating a more uniform dose with the FIF compensation. All calculated monitor units were well within clinically acceptable limits and exceeded those of the original lead compensation plan by less than 50 MU (only ~1.1% increase). Conclusion: MLC FIF technique for TBI will not significantly increase the beam on time while it can substantially reduce the compensator setup time and the potential risk of errors in manually placing lead compensators.

  19. Megavoltage Image-Based Dynamic Multileaf Collimator Tracking of a NiTi Stent in Porcine Lungs on a Linear Accelerator

    SciTech Connect

    Poulsen, Per R.; Carl, Jesper; Nielsen, Jane; Nielsen, Martin S.; Thomsen, Jakob B.; Jensen, Henrik K.; Kjaergaard, Benedict; Zepernick, Peter R.; Worm, Esben; Fledelius, Walther; Cho, Byungchul; Sawant, Amit; Ruan, Dan; Keall, Paul J.

    2012-02-01

    Purpose: To investigate the accuracy and potential limitations of MV image-based dynamic multileaf collimator (DMLC) tracking in a porcine model on a linear accelerator. Methods and Materials: A thermo-expandable NiTi stent designed for kilovoltage (kV) X-ray visualization of lung lesions was inserted into the bronchia of three anaesthetized Goettingen minipigs. A four-dimensional computed tomography scan was used for planning a five-field conformal treatment with circular multileaf collimator (MLC) apertures. A 22.5 Gy single fraction treatment was delivered to the pigs. The peak-to-peak stent motion was 3 to 8 mm, with breathing periods of 1.2 to 4 s. Before treatment, X-ray images were used for image-guided setup based on the stent. During treatment delivery, continuous megavoltage (MV) portal images were acquired at 7.5 Hz. The stent was segmented in the images and used for continuous adaptation of the MLC aperture. Offline, the tracking error in beam's eye view of the treatment beam was calculated for each MV image as the difference between the MLC aperture center and the segmented stent position. The standard deviations of the systematic error {Sigma} and the random error {sigma} were determined and compared with the would-be errors for a nontracking treatment with pretreatment image-guided setup. Results: Reliable stent segmentation was obtained for 11 of 15 fields. Segmentation failures occurred when image contrast was dominated by overlapping anatomical structures (ribs, diaphragm) rather than by the stent, which was designed for kV rather than MV X-ray visibility. For the 11 fields with reliable segmentation, {Sigma} was 0.5 mm/0.4 mm in the two imager directions, whereas {sigma} was 0.5 mm/1.1 mm. Without tracking, {Sigma} and {sigma} would have been 1.7 mm/1.4 mm and 0.8 mm/1.4 mm, respectively. Conclusion: For the first time, in vivo DMLC tracking has been demonstrated on a linear accelerator showing the potential for improved targeting accuracy. The

  20. Commissioning and quality assurance for intensity modulated radiotherapy with dynamic multileaf collimator: experience of the Pontificia Universidad Católica de Chile.

    PubMed

    Venencia, Carlos Daniel; Besa, Pelayo

    2004-01-01

    The objective of this paper is to present our experience in the commissioning and quality assurance (QA) for intensity modulated radiotherapy (IMRT) using dynamic multileaf collimator (dMLC), sliding window technique. Using Varian equipment solution, the connectivity and operation between all IMRT chain components was checked. Then the following test were done: stability of leaf positioning and leaf speed, sensitivity to treatment interruptions (acceleration and deceleration), evaluation of standard field patterns, stability of dMLC output, segmental dose accuracy check, average leaf transmission, dosimetric leaf separation, effects of lateral disequilibrium between adjacent leaves in dose profiles and multiple carriage field verification. Standard patterns were generated for verification: uniform field, pyramid, hole, wedge, peaks and chair. Weekly QA Protocol include: sweeping gap output, Garden Fence Test (narrow bands, 2 mm wide, of exposure spaced at 2-cm intervals) and segmental dose accuracy check. Monthly QA include: sweeping gap output at multiple gantry and collimator angle, sweeping gap output off-axis, Picket Fence Test (eight consecutive movements of a 5-cm wide rectangular field spaced at 5-cm intervals), stability of leaf speed and leaf motor current test (PWM test). Patient QA procedure consists of an absolute dose measurement for all treatments fields in the treatment condition, analysis of actual leaf position versus planned leaf position (dynalog files) for each treatment field, film relative dose determination for each field, film relative dose determination for the plan (all treatment fields) in two axial planes and patient positioning verification with orthogonal films. The tests performed showed acceptable result. After more than one year of IMRT treatment the routine QA machine checks confirm the precision and stability of the IMRT system.

  1. Experimental demonstration of self-collimation inside a three-dimensional photonic crystal.

    PubMed

    Lu, Zhaolin; Shi, Shouyuan; Murakowski, Janusz A; Schneider, Garrett J; Schuetz, Christopher A; Prather, Dennis W

    2006-05-01

    We present our experimental demonstration of self-collimation inside a three-dimensional (3D) simple cubic photonic crystal at microwave frequencies. The photonic crystal was designed with unique dispersion property and fabricated by a high precision computer-controlled machine. The self-collimation modes were excited by a grounded waveguide feeding and detected by a scanning monopole. Self-collimation of electromagnetic waves in the 3D photonic crystal was demonstrated by measuring the 3D field distribution, which was shown as a narrow collimated beam inside the 3D photonic crystal but a diverged beam in the absence of the photonic crystal. PMID:16712297

  2. Self-collimation-based photonic crystal Mach-Zehnder demultiplexer

    NASA Astrophysics Data System (ADS)

    Lee, Sun-Goo; Jung, Soo-Yong; Lee, Jongjin; Park, Jong-Moon; Kee, Chul-Sik

    2016-09-01

    A photonic crystal Mach-Zehnder demultiplexer (PC-MZDmux) with four output ports based on the self-collimation phenomenon in a two-dimensional (2D) PC is proposed and numerically studied using finite-difference time-domain simulations. The PC-MZDmux is composed of three Mach-Zehnder interferometers (MZIs) and each MZI consists of two 50:50 beam splitters and two perfect mirrors. Employed as the design parameters to achieve the demultiplexing functionality are the radius of phase control rods (PCRs) in the mirrors and the distance between the beam spitter and the mirror in the three MZIs. From spatial electric field distributions and transmission spectra, it is demonstrated that an incident self-collimated beam with four different frequencies can be demultiplexed to four output ports of the PC-MZDmux with proper design parameters. Our results indicate that this device design may constitute an efficient approach to light propagation manipulation and increase the application range of self-collimated beams.

  3. Self-collimation-based photonic crystal Mach–Zehnder demultiplexer

    NASA Astrophysics Data System (ADS)

    Lee, Sun-Goo; Jung, Soo-Yong; Lee, Jongjin; Park, Jong-Moon; Kee, Chul-Sik

    2016-09-01

    A photonic crystal Mach–Zehnder demultiplexer (PC-MZDmux) with four output ports based on the self-collimation phenomenon in a two-dimensional (2D) PC is proposed and numerically studied using finite-difference time-domain simulations. The PC-MZDmux is composed of three Mach–Zehnder interferometers (MZIs) and each MZI consists of two 50:50 beam splitters and two perfect mirrors. Employed as the design parameters to achieve the demultiplexing functionality are the radius of phase control rods (PCRs) in the mirrors and the distance between the beam spitter and the mirror in the three MZIs. From spatial electric field distributions and transmission spectra, it is demonstrated that an incident self-collimated beam with four different frequencies can be demultiplexed to four output ports of the PC-MZDmux with proper design parameters. Our results indicate that this device design may constitute an efficient approach to light propagation manipulation and increase the application range of self-collimated beams.

  4. Dosimetric impact of geometric errors due to respiratory motion prediction on dynamic multileaf collimator-based four-dimensional radiation delivery

    SciTech Connect

    Vedam, S.; Docef, A.; Fix, M.; Murphy, M.; Keall, P.

    2005-06-15

    The synchronization of dynamic multileaf collimator (DMLC) response with respiratory motion is critical to ensure the accuracy of DMLC-based four dimensional (4D) radiation delivery. In practice, however, a finite time delay (response time) between the acquisition of tumor position and multileaf collimator response necessitates predictive models of respiratory tumor motion to synchronize radiation delivery. Predicting a complex process such as respiratory motion introduces geometric errors, which have been reported in several publications. However, the dosimetric effect of such errors on 4D radiation delivery has not yet been investigated. Thus, our aim in this work was to quantify the dosimetric effects of geometric error due to prediction under several different conditions. Conformal and intensity modulated radiation therapy (IMRT) plans for a lung patient were generated for anterior-posterior/posterior-anterior (AP/PA) beam arrangements at 6 and 18 MV energies to provide planned dose distributions. Respiratory motion data was obtained from 60 diaphragm-motion fluoroscopy recordings from five patients. A linear adaptive filter was employed to predict the tumor position. The geometric error of prediction was defined as the absolute difference between predicted and actual positions at each diaphragm position. Distributions of geometric error of prediction were obtained for all of the respiratory motion data. Planned dose distributions were then convolved with distributions for the geometric error of prediction to obtain convolved dose distributions. The dosimetric effect of such geometric errors was determined as a function of several variables: response time (0-0.6 s), beam energy (6/18 MV), treatment delivery (3D/4D), treatment type (conformal/IMRT), beam direction (AP/PA), and breathing training type (free breathing/audio instruction/visual feedback). Dose difference and distance-to-agreement analysis was employed to quantify results. Based on our data, the

  5. Intensity modulated radiation therapy with irregular multileaf collimated field: A dosimetric study on the penumbra region with different leaf stepping patterns

    SciTech Connect

    Chow, James C. L.; Grigorov, Grigor N.; Jiang Runqing

    2006-12-15

    Using a Varian 21 EX linear accelerator with a multileaf collimator (MLC) of 120 leaves, the penumbra regions of beam profiles within an irregular multileaf collimated fields were studied. MLC fields with different leaf stepping angles from 21.8 deg. to 68.2 deg. were used. Beam profiles in different directions: (1) along the cross-line and in-line axis (2) along the leaf stepping edges of the field, and (3) parallel to the stepping edges but in the middle of the field, were measured and calculated using Kodak XV radiographic film and Pinnacle3 treatment planning system version 7.4f. These beam profiles were measured and calculated at source to axis distance=100 cm with 5 cm of solid water slab on top. On the one hand, for both cross-line and in-line beam profiles, the penumbra widths of 20%-80% did not vary with the leaf stepping angles and were about 0.4 cm. On the other hand, the penumbra widths of 10%-90% of the above two profiles varied with the stepping angles and had maximum widths of about 1.9 cm (cross-line) and 1.65 cm (in-line) for stepping angles of 38.7 deg. and 51.3 deg., respectively. For profiles crossing the 'rippled' stepping edges of the field, the penumbra widths (10%-90%) at the regions between two opposite leaves (i.e., profile end at the Y1/Y2 jaw position) decreased with the stepping angles. At the penumbra regions between two leaf edges with the tongue-and-groove structure of the same bank (i.e., profile end at the X1/X2 jaw position), the penumbra widths increased with the stepping angles. When the penumbra widths were measured between two opposite leaf edges and at corners between two leaves, the widths first decreased with the stepping angles and then increased beyond the minimum width point at stepping angle of 45 deg. The penumbra width (10%-90%) measured at the leaf edge was larger than that at the corner. For the beam profiles calculated using Pinnacle3, it is found that the results agreed well with the measurements along the cross

  6. Integration of Real-Time Internal Electromagnetic Position Monitoring Coupled With Dynamic Multileaf Collimator Tracking: An Intensity-Modulated Radiation Therapy Feasibility Study

    SciTech Connect

    Smith, Ryan L.; Sawant, Amit PhD.; Santanam, Lakshmi PhD.; Venkat, Raghu B.; Newell, Laurence J.; Cho, Byung-chul; Poulsen, Per; Catell, Herbert; Keall, Paul J.; Parikh, Parag J.

    2009-07-01

    Purpose: Continuous tumor position measurement coupled with a tumor tracking system would result in a highly accurate radiation therapy system. Previous internal position monitoring systems have been limited by fluoroscopic radiation dose and low delivery efficiency. We aimed to incorporate a continuous, electromagnetic, three-dimensional position tracking system (Calypso 4D Localization System) with a dynamic multileaf collimator (DMLC)-based dose delivery system. Methods and Materials: A research version of the Calypso System provided real-time position of three Beacon transponders. These real-time three-dimensional positions were sent to research MLC controller with a motion-tracking algorithm that changed the planned leaf sequence. Electromagnetic transponders were embedded in a solid water film phantom that moved with patient lung trajectories while being irradiated with two different plans: a step-and-shoot intensity-modulated radiation therapy (S-IMRT) field and a dynamic IMRT (D-IMRT) field. Dosimetric results were recorded under three conditions: no intervention, DMLC tracking, and a spatial gating system. Results: Dosimetric accuracy was comparable for gating and DMLC tracking. Failure rates for gating/DMLC tracking are as follows: {+-}3 cGy 10.9/ 7.5% for S-IMRT, 3.3/7.2% for D-IMRT; gamma (3mm/3%) 0.2/1.2% for S-IMRT, 0.2/0.2% for D-IMRT. DMLC tracking proved to be as efficient as standard delivery, with a two- to fivefold efficiency increase over gating. Conclusions: Real-time target position information was successfully integrated into a DMLC effector system to modify dose delivery. Experimental results show both comparable dosimetric accuracy as well as improved efficiency compared with spatial gating.

  7. SU-E-T-424: Dosimetric Verification of Modulated Electron Radiation Therapy Delivered Using An Electron Specific Multileaf Collimator for Treatment of Scalp Cases

    SciTech Connect

    Eldib, A; Jin, L; Martin, J; Li, J; Chibani, O; Galloway, T; Ma, C; Mora, G

    2014-06-01

    Purpose: Modulated electron radiotherapy (MERT) has the potential to achieve better treatment outcome for shallow tumors such as those of breast and scalp. In a separate study with scalp lesions, MERT was compared to volumetric modulated arc therapy. Our results showed a reduction in the dose reaching the brain with MERT. However dose calculation accuracy and delivery efficiency challenges remain. Thus in the current study we proceed to add more cases to demonstrate MERT beneficial outcome and its delivery accuracy using an electron specific multileaf collimator (eMLC). Methods: We have used the MCBEAM code for treatment head simulation and for generating phase space files to be used as radiation source input for our Monte Carlo based treatment planning system (MC TPS). MCPLAN code is used for calculation of patient specific dose deposition coefficient and for final MERT plan dose calculation. An in-house developed optimization code is used for the optimization process. MERT plans were generated for real patients and head and neck phantom. Film was used for dosimetric verification. The film was cut following the contour of the curved phantom surface and then sealed with black masking tape. In the measurement, the sealed film packet was sandwiched between two adjacent slabs of the head and neck phantom. The measured 2D dose distribution was then compared with calculations. Results: The eMLC allows effective treatment of scalps with multi-lesions spreading around the patient head, which was usually difficult to plan or very time consuming with conventional applicators. MERT continues to show better reduction in the brain dose. The dosimetric measurements showed slight discrepancy, which was attributed to the film setup. Conclusion: MERT can improve treatment plan quality for patients with scalp cancers. Our in-house MC TPS is capable of performing treatment planning and accurate dose calculation for MERT using the eMLC.

  8. Impact of the number of control points has on isodose distributions in a dynamic multileaf collimator intensity-modulated radiation therapy delivery

    SciTech Connect

    Goraj, Andrew; Boer, Steven F. de

    2012-01-01

    Intensity-modulated radiation therapy (IMRT) is a powerful technique in planning the delivery of dose. The most common IMRT delivery requires the use of moving multileaf collimators (MLCs) to deliver the requested fluence pattern. A dynamic delivery IMRT field file will contain several control points that are defined MLC shapes at a marked fraction of the delivered monitor units. The size of this file and the fidelity of the deliverable fluence are proportional to the number of control points defined. This study investigates the effect of reducing the number of control points has on the resultant dose distribution quality in complex IMRT in efforts to reduce transfer times, loading times, check sum times and file storage. Analysis was performed with 6 head and neck patients on an Eclipse version 8.5 treatment planning system (Varian, Palo Alto, CA). To ensure the quality of all treatments, Eclipse defines a minimum of 64 and a maximum of 320 control points per subfield (Eclipse Algorithms Reference guide). All 6 patients' plans were calculated with fixed 64, 166, and 320 control points using the sliding window technique. In addition, each plan was calculated in variable mode (Normal mode) in which the planning system determined the required number of control points. Each of the 4 plans for each patient was renormalized to provide the same mean planning target volume (PTV) 70 dose. Dose values for critical and target structures were examined for each patient. When examining the minimum, maximum, and mean doses to all target structures, it was noted that the greatest reduction in target dose coverage caused by reduced number of control points was 0.5%, which occurred for the minimum dose to the PTV56 structure in one plan.' Dose analysis for critical structures showed no clinically significant increase in dose when compared with the 320 control point plan.

  9. Bending self-collimated one-way light by using gyromagnetic photonic crystals

    SciTech Connect

    Li, Qing-Bo; Li, Zhen; Wu, Rui-xin

    2015-12-14

    We theoretically demonstrate that electromagnetic waves can self-collimate and propagate unidirectionally in photonic crystals fabricated using semicylindrical ferrite rods in magnetized states. The parity and time-reversal symmetries of such photonic crystals are broken, resulting in a self-collimated one-way body wave within the photonic crystals. By applying the bias magnetic field in a complex configuration, the self-collimated one-way wave beam can be bent into arbitrary trajectories within the photonic crystal, providing an avenue for controlling wave beams.

  10. Feasibility of using Geant4 Monte Carlo simulation for IMRT dose calculations for the Novalis Tx with a HD-120 multi-leaf collimator

    NASA Astrophysics Data System (ADS)

    Jung, Hyunuk; Shin, Jungsuk; Chung, Kwangzoo; Han, Youngyih; Kim, Jinsung; Choi, Doo Ho

    2015-05-01

    The aim of this study was to develop an independent dose verification system by using a Monte Carlo (MC) calculation method for intensity modulated radiation therapy (IMRT) conducted by using a Varian Novalis Tx (Varian Medical Systems, Palo Alto, CA, USA) equipped with a highdefinition multi-leaf collimator (HD-120 MLC). The Geant4 framework was used to implement a dose calculation system that accurately predicted the delivered dose. For this purpose, the Novalis Tx Linac head was modeled according to the specifications acquired from the manufacturer. Subsequently, MC simulations were performed by varying the mean energy, energy spread, and electron spot radius to determine optimum values of irradiation with 6-MV X-ray beams by using the Novalis Tx system. Computed percentage depth dose curves (PDDs) and lateral profiles were compared to the measurements obtained by using an ionization chamber (CC13). To validate the IMRT simulation by using the MC model we developed, we calculated a simple IMRT field and compared the result with the EBT3 film measurements in a water-equivalent solid phantom. Clinical cases, such as prostate cancer treatment plans, were then selected, and MC simulations were performed. The accuracy of the simulation was assessed against the EBT3 film measurements by using a gamma-index criterion. The optimal MC model parameters to specify the beam characteristics were a 6.8-MeV mean energy, a 0.5-MeV energy spread, and a 3-mm electron radius. The accuracy of these parameters was determined by comparison of MC simulations with measurements. The PDDs and the lateral profiles of the MC simulation deviated from the measurements by 1% and 2%, respectively, on average. The computed simple MLC fields agreed with the EBT3 measurements with a 95% passing rate with 3%/3-mm gamma-index criterion. Additionally, in applying our model to clinical IMRT plans, we found that the MC calculations and the EBT3 measurements agreed well with a passing rate of greater

  11. Tracking 'differential organ motion' with a 'breathing' multileaf collimator: magnitude of problem assessed using 4D CT data and a motion-compensation strategy

    NASA Astrophysics Data System (ADS)

    McClelland, J. R.; Webb, S.; McQuaid, D.; Binnie, D. M.; Hawkes, D. J.

    2007-08-01

    Intrafraction tumour (e.g. lung) motion due to breathing can, in principle, be compensated for by applying identical breathing motions to the leaves of a multileaf collimator (MLC) as intensity-modulated radiation therapy is delivered by the dynamic MLC (DMLC) technique. A difficulty arising, however, is that irradiated voxels, which are in line with a bixel at one breathing phase (at which the treatment plan has been made), may move such that they cease to be in line with that breathing bixel at another phase. This is the phenomenon of differential voxel motion and existing tracking solutions have ignored this very real problem. There is absolutely no tracking solution to the problem of compensating for differential voxel motion. However, there is a strategy that can be applied in which the leaf breathing is determined to minimize the geometrical mismatch in a least-squares sense in irradiating differentially-moving voxels. A 1D formulation in very restricted circumstances is already in the literature and has been applied to some model breathing situations which can be studied analytically. These are, however, highly artificial. This paper presents the general 2D formulation of the problem including allowing different importance factors to be applied to planning target volume and organ at risk (or most generally) each voxel. The strategy also extends the literature strategy to the situation where the number of voxels connecting to a bixel is a variable. Additionally the phenomenon of 'cross-leaf-track/channel' voxel motion is formally addressed. The general equations are presented and analytic results are given for some 1D, artificially contrived, motions based on the Lujan equations of breathing motion. Further to this, 3D clinical voxel motion data have been extracted from 4D CT measurements to both assess the magnitude of the problem of 2D motion perpendicular to the beam-delivery axis in clinical practice and also to find the 2D optimum breathing-leaf strategy

  12. SU-E-T-428: Dosimetric Impact of Multileaf Collimator Leaf Width On Single and multiple Isocenter Stereotactic IMRT Treatment Plans for multiple Brain Tumors

    SciTech Connect

    Giem, J; Algan, O; Ahmad, S; Ali, I; Young, J; Hossain, S

    2014-06-01

    Purpose: To assess the impacts that multileaf collimator (MLC) leaf width has on the dose conformity and normal brain tissue doses of single and multiple isocenter stereotactic IMRT (SRT) plans for multiple intracranial tumors. Methods: Fourteen patients with 2–3 targets were studied retrospectively. Patients treated with multiple isocenter treatment plans using 9 to 12 non-coplanar beams per lesion underwent repeat planning using single isocenter and 10 to 12 non-coplanar beams with 2.5mm, 3mm and 5mm MLC leaf widths. Brainlab iPlan treatment planning system for delivery with the 2.5mm MLC served as reference. Identical contour sets and dose-volume constraints were applied. The prescribed dose to each target was 25 Gy to be delivered over 5 fractions with a minimum of 99% dose to cover ≥ 95% of the target volume. Results: The lesions and normal brains ranged in size from 0.11 to 51.67cc (median, 2.75cc) and 1090 to 1641cc (median, 1401cc), respectively. The Paddick conformity index for single and multiple isocenter (2.5mm vs. 3mm and 5mm MLCs) was (0.79±0.08 vs. 0.79±0.07 and 0.77±0.08) and (0.79±0.09 vs. 0.77±0.09 and 0.76±0.08), respectively. The average normal brain volumes receiving 15 Gy for single and multiple isocenter (2.5mm vs. 3mm and 5mm MLCs) were (3.65% vs. 3.95% and 4.09%) and (2.89% vs. 2.91% and 2.92%), respectively. Conclusion: The average dose conformity observed for the different leaf width for single and multiple isocenter plans were similar, throughout. However, the average normal brain volumes receiving 2.5 to 15 Gy were consistently lower for the 2.5mm MLC leaf width, especially for single isocenter plans. The clinical consequences of these integral normal brain tissue doses are still unknown, but employing the use of the 2.5mm MLC option is desirable at sparing normal brain tissue for both single and multiple isocenter cases.

  13. Discussion on the usefulness of dose dynamic multi-leaf collimator-based plan to overcome dose limit of spinal cord in high-dose radiotherapy

    NASA Astrophysics Data System (ADS)

    Kim, E. C.; Cho, J. H.; Park, C. S.; Kim, D. H.; Choi, C. W.

    2014-03-01

    In this study, the conventional plan was compared with the plan that was based on a dose dynamic multi-leaf collimator (MLC), and a dose dynamic MLC was used to evaluate its usefulness. Then, this study examined if it was possible to perform a high-dose radiation therapy by adjusting the dose limit of the spinal cord when the dose dynamic MLC-based plan was used. First of all, linear accelerator was used to compare the conventional plan with the dose dynamic MLC-based plan. Then, the study was conducted in two methods in order to examine the proper range of the shield for the spinal cord when the dose dynamic MLC was used to adjust the dose of the spinal cord. In the first method, X-omat film was used to perform film dosimetry. In the second method, radiation treatment planning (RTP) system was used to find out the proper range among 0, 3, 6, and 9 mm. When film scan was performed in the each range, respectively, from the spinal cord and under the same conditions, it was confirmed to be appropriate to use the range of 3 mm. When the RTP system was used to perform planning in the shield range of each range, respectively, from the spinal cord, dose-volume histogram (DVH) was a slight difference could be found in the region from 25% to 35%. On the contrary, no radiation exposure was found in the region of 35% or higher for all of the each range. Consequently, if MLC is selected in consideration of the planning target volume (PTV), the most proper value can be obtained by selecting the range of 3 mm. Next, the DVH was compared to examine the relationship in PTV when the each range was used for planning. All of the ranges showed the same pattern up to the point of 90%. However, the results were different in the region of higher than 90% because the dose was low for the spinal cord, and a relatively useful dose was used for PTV when the range was 3 mm.

  14. Photonic drop splitters based on silicon photonic crystal cascaded self-collimation ring resonators

    NASA Astrophysics Data System (ADS)

    Lin, Yuan-Yuan; Chen, Xi-Yao; Jiang, Jun-Zhen; Fu, Ping; Chen, Biao; Yang, Li-Hui; Liu, Jing-Ping; Lin, Bao-Cheng

    2014-10-01

    In this paper, the 1×5 optical splitters (OSs) based on 2D rod-type silicon photonic crystal embed cascaded self-collimation (SC) ring resonators (CSCRR) was proposed. The 1×5 OSs consist of eight beam splitters, which are formed by varying the radii of the rod. With self-collimation effect, we can manipulate the light's propagation in the OSs. Here we consider TM modes. Utilizing multiple-beam interference theory, the theoretical transmission spectra at different outputs were analysed. These transmission spectra can help us to set the radii of eight slitters properly, for we can control the light coming out from five ports with the light-intensity ratio we need. Meanwhile these outputs' transmission spectra were investigated by the finite-difference time-domain (FDTD) method. The simulative results have an agreement with the theoretical prediction. The 1×5 OSs will have practical applications in photonic integrated circuits.

  15. Electronic tissue compensation achieved with both dynamic and static multileaf collimator in eclipse treatment planning system for Clinac 6 EX and 2100 CD Varian linear accelerators: Feasibility and dosimetric study

    PubMed Central

    Kinhikar, Rajesh A.; Sharma, Pramod K.; Patkar, Sachin; Tambe, Chandrashekhar M.; Deshpande, Deepak D.

    2007-01-01

    Dynamic multileaf collimator (DMLC) and static multileaf collimator (SMLC), along with three-dimensional treatment planning system (3-D TPS), open the possibility of tissue compensation. A method using electronic tissue compensator (ETC) has been implemented in Eclipse 3-D TPS (V 7.3, Varian Medical Systems, Palo Alto, USA) at our center. The ETC was tested for head and neck conformal radiotherapy planning. The purpose of this study was to verify the feasibility of DMLC and SMLC in head and neck field irradiation for delivering homogeneous dose in the midplane at a pre-defined depth. In addition, emphasis was given to the dosimetric aspects in commissioning ETC in Eclipse. A Head and Neck Phantom (The Phantom Laboratory, USA) was used for the dosimetric verification. Planning was carried out for both DMLC and SMLC ETC plans. The dose calculated at central axis by eclipse with DMLC and SMLC was noted. This was compared with the doses measured on machine with ion chamber and thermoluminescence dosimetry (TLD). The calculated isodose curves and profiles were compared with the measured ones. The dose profiles along the two major axes from Eclipse were also compared with the profiles obtained from Amorphous Silicon (AS500) Electronic portal imaging device (EPID) on Clinac 6 EX machine. In uniform dose regions, measured dose values agreed with the calculated doses within 3%. Agreement between calculated and measured isodoses in the dose gradient zone was within 3 mm. The isodose curves and the profiles were found to be in good agreement with the measured curves and profiles. The measured and the calculated dose profiles along the two major axes were flat for both DMLC and SMLC. The dosimetric verification of ETC for both the linacs demonstrated the feasibility and the accuracy of the ETC treatment modality for achieving uniform dose distributions. Therefore, ETC can be used as a tool in head and neck treatment planning optimization for improved dose uniformity. PMID

  16. All-angle self-collimation in two-dimensional rhombic-lattice photonic crystals

    NASA Astrophysics Data System (ADS)

    Wu, Z. H.; Xie, K.; Yang, H. J.; Jiang, P.; He, X. J.

    2012-01-01

    In this work, the self-collimating phenomenon in two-dimensional rhombic-lattice photonic crystals is discussed in detail. The equi-frequency contours of a rhombic-lattice rod-type photonic crystal structure are calculated with the plane wave expansion method for different lattice angles and rod radii. A broad range of flat and open equi-frequency contours is found. A rhombic-lattice photonic crystal made of 301 × 21 silicon rods (n = 3.5) in air is presented as an illustration to demonstrate the broadband all-angle self-collimation. The bandwidth for all-angle self-collimation is verified to be 17.42% of the central frequency by simulations based on the finite-difference time-domain method. The all-angle self-collimation is found to be polarization-independent and the frequency for simultaneous TM and TE all-angle self-collimation is identified and verified by finite-difference time-domain simulations.

  17. Direct impact analysis of multi-leaf collimator leaf position errors on dose distributions in volumetric modulated arc therapy: a pass rate calculation between measured planar doses with and without the position errors

    NASA Astrophysics Data System (ADS)

    Tatsumi, D.; Hosono, M. N.; Nakada, R.; Ishii, K.; Tsutsumi, S.; Inoue, M.; Ichida, T.; Miki, Y.

    2011-10-01

    We propose a new method for analyzing the direct impact of multi-leaf collimator (MLC) leaf position errors on dose distributions in volumetric modulated arc therapy (VMAT). The technique makes use of the following processes. Systematic leaf position errors are generated by directly changing a leaf offset in a linac controller; dose distributions are measured by a two-dimensional diode array; pass rates of the dose difference between measured planar doses with and without the position errors are calculated as a function of the leaf position error. Three different treatment planning systems (TPSs) were employed to create VMAT plans for five prostate cancer cases and the pass rates were compared between the TPSs under various leaf position errors. The impact of the leaf position errors on dose distributions depended upon the final optimization result from each TPS, which was explained by the correlation between the dose error and the average leaf gap width. The presented method determines leaf position tolerances for VMAT delivery for each TPS, which may facilitate establishing a VMAT quality assurance program in a radiotherapy facility. This work was presented in part at the 52nd Annual Meeting of the American Society for Therapeutic Radiology and Oncology in San Diego on 1 November 2010.

  18. SU-E-T-604: Penumbra Characteristics of a New InCiseâ„¢ Multileaf Collimator of CyberKnife M6â„¢ System

    SciTech Connect

    Hwang, M; Jang, S; Ozhasoglu, C; Lalonde, R; Heron, D; Huq, M

    2015-06-15

    Purpose: The InCise™ Multileaf Collimator (MLC) of CyberKnife M6™ System has been released recently. The purpose of this study was to explore the dosimetric characteristics of the new MLC. In particular, the penumbra characteristics of MLC fields at varying locations are evaluated. Methods: EBT3-based film measurements were performed with varying MLC fields ranging from 7.5 mm to 27.5 mm. Seventeen regions of interests (ROIs) were identified for irradiation. These are regions located at the central area (denoted as reference field), at the left/right edge areas of reference open field, at an intermediate location between central and edge area. Single beam treatment plans were designed by using the MultiPlan and was delivered using the Blue Phantom. Gafchromic films were irradiated at 1.5 cm depth in the Blue Phantom and analyzed using the Film Pro software. Variation of maximum dose, penumbra of MLC-defined fields, and symmetry/flatness were calculated as a function of locations of MLC fields. Results: The InCise™ MLC System showed relatively consistent dose distribution and penumbra size with varying locations of MLC fields. The measured maximum dose varied within 5 % at different locations compared to that at the central location and agreed with the calculated data well within 2%. The measured penumbrae were in the range of 2.9 mm and 3.7 mm and were relatively consistent regardless of locations. However, dose profiles in the out-of-field and in-field regions varied with locations and field sizes. Strong variation was seen for all fields located at 55 mm away from the central field. The MLC leakage map showed that the leakage is dependent on position. Conclusion: The size of penumbra and normalized maximum dose for MLC-defined fields were consistent in different regions of MLC. However, dose profiles in the out-field region varied with locations and field sizes.

  19. Feasibility of single-isocenter, multi-arc non-coplanar volumetric modulated arc therapy for multiple brain tumors using a linear accelerator with a 160-leaf multileaf collimator: a phantom study

    PubMed Central

    Iwai, Yoshio; Ozawa, Shuichi; Ageishi, Tatsuya; Pellegrini, Roberto; Yoda, Kiyoshi

    2014-01-01

    The feasibility of single isocenter, multi-arc non-coplanar volumetric modulated arc therapy (VMAT) for multiple brain tumors was studied using an Elekta Synergy linear accelerator with an Agility multileaf collimator and a Monaco treatment planning system. Two VMAT radiosurgery plans consisting of a full arc and three half arcs were created with a prescribed dose of 20 Gy in a single fraction. After dose delivery to a phantom, ionization chambers and radiochromic films were used for dose measurement. The first VMAT radiosurgery plan had nine targets inside the phantom, and the doses were measured by the chambers at two different points and by the films on three sagittal and three coronal planes. The differences between the calculated dose and the dose measured by a Farmer ionization chamber and a pinpoint ionization chamber were <1.00% and <2.30%, respectively, and the average pass rates of gamma indices among the six planes under each of 3%/3 mm and 2%/2 mm criteria were 98.6% and 92.6%, respectively. The second VMAT radiosurgery plan was based on a clinical 14 brain metastases. Differences between calculated and film-measured doses were evaluated on two sagittal planes. The average pass rates of the gamma indices on the planes under each of 3%/3 mm and 2%/2 mm criteria were 97.8% and 88.8%, respectively. It was confirmed that single-isocenter, non-coplanar multi-arc VMAT radiosurgery for multiple brain metastases was feasible using Elekta Synergy with Agility and Monaco treatment planning systems. It was further shown that film dosimetry was accurately performed for a dose of up to nearly 25 Gy. PMID:24944266

  20. Feasibility of single-isocenter, multi-arc non-coplanar volumetric modulated arc therapy for multiple brain tumors using a linear accelerator with a 160-leaf multileaf collimator: a phantom study.

    PubMed

    Iwai, Yoshio; Ozawa, Shuichi; Ageishi, Tatsuya; Pellegrini, Roberto; Yoda, Kiyoshi

    2014-09-01

    The feasibility of single isocenter, multi-arc non-coplanar volumetric modulated arc therapy (VMAT) for multiple brain tumors was studied using an Elekta Synergy linear accelerator with an Agility multileaf collimator and a Monaco treatment planning system. Two VMAT radiosurgery plans consisting of a full arc and three half arcs were created with a prescribed dose of 20 Gy in a single fraction. After dose delivery to a phantom, ionization chambers and radiochromic films were used for dose measurement. The first VMAT radiosurgery plan had nine targets inside the phantom, and the doses were measured by the chambers at two different points and by the films on three sagittal and three coronal planes. The differences between the calculated dose and the dose measured by a Farmer ionization chamber and a pinpoint ionization chamber were <1.00% and <2.30%, respectively, and the average pass rates of gamma indices among the six planes under each of 3%/3 mm and 2%/2 mm criteria were 98.6% and 92.6%, respectively. The second VMAT radiosurgery plan was based on a clinical 14 brain metastases. Differences between calculated and film-measured doses were evaluated on two sagittal planes. The average pass rates of the gamma indices on the planes under each of 3%/3 mm and 2%/2 mm criteria were 97.8% and 88.8%, respectively. It was confirmed that single-isocenter, non-coplanar multi-arc VMAT radiosurgery for multiple brain metastases was feasible using Elekta Synergy with Agility and Monaco treatment planning systems. It was further shown that film dosimetry was accurately performed for a dose of up to nearly 25 Gy.

  1. Small animal imaging by single photon emission using pinhole and coded aperture collimation

    SciTech Connect

    Garibaldi, F.; Accorsi, R.; Cinti, M.N.; Colilli, S.; Cusanno, F.; De Vincentis, G.; Fortuna, A.; Girolami, B.; Giuliani, F.; Gricia, M.; Lanza, R.; Loizzo, A.; Loizzo, S.; Lucentini, M.; Majewski, S.; Santavenere, F.; Pani, R.; Pellegrini, R.; Signore, A.; Scopinaro, F.

    2005-06-01

    The aim of this paper is to investigate the basic properties and limits of the small animal imaging systems based on single photon detectors. The detectors for radio imaging of small animals are challenging because of the very high spatial resolution needed, possibly coupled with high efficiency to allow dynamic studies. These performances are hardly attainable with single photon technique because of the collimator that limits both spatial resolution and sensitivity. In this paper we describe a simple desktop detector based on pixellated NaI(Tl) scintillator array coupled with a pinhole collimator and a PSPMT, the Hamamatsu R2486. The limits of such systems as well as the way to overcome them will be shown. In fact better light sampling at the anode level would allow better pixel identification for higher number of pixel that is one of the parameters defining the image quality. Also the spatial resolution would improve. The performances of such layout are compared with others using PSPMTs differing from R2486 for the light sampling at the anode level and different areas. We show how a further step, namely the substitution of the pinhole collimator with a coded aperture, will allow a great improvement in system sensitivity while maintaining very good spatial resolution, possibly submillimetric. Calculations and simulations show that sensitivity would improve by a factor of 50.

  2. 2D photonic crystal logic gates based on self-collimated effect

    NASA Astrophysics Data System (ADS)

    Fan, Ranran; Yang, Xiulun; Meng, Xiangfeng; Sun, Xiaowen

    2016-08-01

    Four kinds of logic gates are proposed using interference between the self-collimated beams in photonic crystals, namely NOT, OR, AND and XOR gates, which can be used in the design of photonic integrated circuits. The radius of the splitter and the optical path difference between splitters are adjusted to produce certain phase difference between the reflected and transmitted beams, which may interfere constructively or destructively to realize logical operation. They have high contrast ratios and low power consumption, the extinction ratio between logic 1 and logic 0 for NOT and AND gates can reach 24.7 dB, 30 dB and 12.6 dB for the wavelength used by optical communication (1550 nm), respectively, which makes it potentially applicable for photonic integrated circuits.

  3. A versatile optical junction using photonic band-gap guidance and self collimation

    SciTech Connect

    Gupta, Man Mohan; Medhekar, Sarang

    2014-09-29

    We show that it is possible to design two photonic crystal (PC) structures such that an optical beam of desired wavelength gets guided within the line defect of the first structure (photonic band gap guidance) and the same beam gets guided in the second structure by self-collimation. Using two dimensional simulation of a design made of the combination of these two structures, we propose an optical junction that allows for crossing of two optical signals of same wavelength and same polarization with very low crosstalk. Moreover, the junction can be operated at number of frequencies in a wide range. Crossing of multiple beams with very low cross talk is also possible. The proposed junction should be important in future integrated photonic circuits.

  4. Photon and electron collimator effects on electron output and abutting segments in energy modulated electron therapy

    SciTech Connect

    Olofsson, Lennart; Karlsson, Magnus G.; Karlsson, Mikael

    2005-10-15

    In energy modulated electron therapy a large fraction of the segments will be arranged as abutting segments where inhomogeneities in segment matching regions must be kept as small as possible. Furthermore, the output variation between different segments should be minimized and must in all cases be well predicted. For electron therapy with add-on collimators, both the electron MLC (eMLC) and the photon MLC (xMLC) contribute to these effects when an xMLC tracking technique is utilized to reduce the x-ray induced leakage. Two add-on electron collimator geometries have been analyzed using Monte Carlo simulations: One isocentric eMLC geometry with an isocentric clearance of 35 cm and air or helium in the treatment head, and one conventional proximity geometry with a clearance of 5 cm and air in the treatment head. The electron fluence output for 22.5 MeV electrons is not significantly affected by the xMLC if the shielding margins are larger than 2-3 cm. For small field sizes and 9.6 MeV electrons, the isocentric design with helium in the treatment head or shielding margins larger than 3 cm is needed to avoid a reduced electron output. Dose inhomogeneity in the matching region of electron segments is, in general, small when collimator positions are adjusted to account for divergence in the field. The effect of xMLC tracking on the electron output can be made negligible while still obtaining a substantially reduced x-ray leakage contribution. Collimator scattering effects do not interfere significantly when abutting beam techniques are properly applied.

  5. Polarization-independent drop filters based on photonic crystal self-collimation ring resonators.

    PubMed

    Chen, Xiyao; Qiang, Zexuan; Zhao, Deyin; Li, Hui; Qiu, Yishen; Yang, Weiquan; Zhou, Weidong

    2009-10-26

    We report here a polarization-independent drop filter (PIDF) based on a photonic crystal self-collimation ring resonator (SCRR). Despite of the large birefringence associated with the polarization-dependent dispersion properties, we demonstrate a PIDF based on multiple-beam interference theory and polarization peak matching (PPM) technique. The PIDF performance was also investigated based on finite-difference time-domain (FDTD) technique, with excellent agreement between the theory and the simulation. For the designed drop wavelength of 1550 nm, the polarization-independent free spectral range is about 36.1 nm, which covers the whole optical communication C-band window. The proposed PIDFs are highly desirable for applications in photonic integrated circuits (PICs). PMID:19997202

  6. Design and performance of a compact collimator on GM/CA-CAT at the Advanced Photon Source

    NASA Astrophysics Data System (ADS)

    Xu, S.; Fischetti, R. F.

    2007-09-01

    A new macromolecular crystallographic facility developed by The General Medicine and Cancer Institutes Collaborative Access Team (GM/CA-CAT) at the Advanced Photon Source (APS) is a part of the Biosciences Division (BIO), Argonne National Laboratory (ANL). The facility consists of three beamlines: two lines based on the first "hard" dual canted undulators and one bending magnet beamline [1]. Several compact collimator systems have been developed for the purpose of background reduction in macromolecular crystallography experiments. The apparatus consists of a tube collimator, pinhole and kinematics mount. This paper will present a series of compact collimator designs and crystallographic applications based on experimental requirements [2]. We also describe the magnet-based kinematic mounting structures [3] developed as a collimator holder.

  7. Experimental two-dimensional field mapping of total internal reflection lateral beam shift in a self-collimated photonic crystal

    NASA Astrophysics Data System (ADS)

    Garcia-Pomar, J. L.; Gollub, J. N.; Mock, J. J.; Smith, D. R.; Nieto-Vesperinas, M.

    2009-02-01

    A lateral beam shift is demonstrated both theoretically and in microwave experiments when total internal reflection takes place at the boundary of a self-collimating two-dimensional photonic crystal consisting of an array of high index dielectric cylinders. We further show the dependence of this shift on the cut of the last row of cylinders that defines the crystal interface.

  8. A 1 × 4 optical splitter for TE modes based on a silicon photonic crystal self-collimation ring resonator

    NASA Astrophysics Data System (ADS)

    Zhong, Yuangang; Chen, Xiyao; Qiang, Zexuan; Lin, Guimin; Jiang, Junzhen; Qiu, Yishen; Li, Hui

    2013-09-01

    A 1 × 4 optical splitter (OS) is proposed for TE modes based on a self-collimation (SC) effect ring resonator (SCRR) in an air-hole type silicon photonic crystal. A 1 × 4 OS consists of four beam splitters formed by varying the radii of the air holes. Utilizing multiple-beam interference theory, the theoretical transmission spectra at each port in the OS were analyzed. By forming four splitters in a SCRR properly, self-collimation light can come out from four ports with the light-intensity ratio we set. OSs were investigated using the two-dimensional finite-difference time-domain (FDTD) simulation technique. The simulation results have good agreement with the theoretical prediction. Because of its small dimensions, whole silicon material, and air-hole type, this structure may have an important role in photonic integrated circuits.

  9. All-angle self-collimation in two-dimensional square array photonic crystals based on index contrast tailoring

    NASA Astrophysics Data System (ADS)

    Noori, Mina; Soroosh, Mohammad; Baghban, Hamed

    2015-03-01

    A comprehensive study has been performed to achieve all-angle self-collimation in basic two-dimensional square array photonic crystals with cylindrical scatterers. Based on plane wave expansion and finite difference time domain analysis for both rod- and hole-type structures, we report on all-angle self-collimation (SC) in the first band of the structure, which results in loss suppression due to out-of-plane scatterings. A lower threshold for index contrast has been obtained to achieve all-angle SC, which offers more design flexibility regarding structural parameters. Furthermore, it has been shown that a minimum and maximum coupling efficiency enhancement of ˜40% and 80% can be achieved for the proposed structure, respectively, by introducing a row of scatterers with proper radius at the input and the output air/photonic crystal interfaces.

  10. UCD-SPI: Un-Collimated Detector Single-Photon Imaging System for Small Animal and Plant Imaging

    NASA Astrophysics Data System (ADS)

    Walker, Katherine Leigh

    Medical imaging systems using single gamma-ray emitting radioisotopes implement collimators in order to form images. However, a tradeoff in sensitivity is inherent in the use of collimators, and modern preclinical single-photon emission computed tomography (SPECT) systems detect a very small fraction of emitted gamma-rays (<0.3%). We have built a collimator-less system, which can reach sensitivity of 40% for 99mTc imaging, while still producing images of sufficient spatial resolution for certain applications in "thin" objects such as mice, small plants, and well plates used for in vitro experiments. This flexible geometry un-collimated detector single-photon imaging (UCD-SPI) system consists of two large (5 cm x 10 cm), thin (3 mm and 5 mm), closely spaced, pixelated scintillation detectors of either NaI(Tl), CsI(Na), or BGO. The detectors are read out by two adjacent Hamamatsu H8500 multichannel photomultiplier tubes. The detector heads enable the interchange of scintillation detectors of different materials and thicknesses to optimize performance for a wide range of gamma-ray energies and imaging subjects. The detectors are horizontally oriented for animal imaging, and for plant imaging the system is rotated on its side to orient the detectors vertically. While this un-collimated detector system is unable to approach the sub-mm spatial resolution obtained by the most advanced preclinical pinhole SPECT systems, the high sensitivity could enable significant and new use in molecular imaging applications which do not require good spatial resolution- for example, screening applications for drug development (small animals), for material transport and sequestration studies for phytoremediation (plants), or for counting radiolabeled cells in vitro (well plates).

  11. The application of Monte Carlo simulation to the design of collimators for single photon emission computed tomography

    NASA Astrophysics Data System (ADS)

    Cullum, Ian Derek

    Single photon emission computed tomography offers the potential for quantification of the uptake of radiopharmaceuticals in-vivo. This thesis investigates some of the factors which limit the accuracy of these methods for measurements in the human brain and investigates how the errors can be reduced. Modifications to data collection devices rather than image reconstruction techniques are studied. To assess the impact of errors on images, a set of computer generated test objects were developed. These included standard Anger and Phelps phantoms and a series of slices of the human brain taken from an atlas of transmission tomography. System design involves a balance between resolution and noise in the image. The optimal resolution depends on the data collection system, the uptake characteristics of the radiopharmaceutical and object size. A method to determine this resolution was developed and showed a single-slice system employing focused, probe detectors to offer greater potential for quantification in the brain than systems based on multiple Anger gamma cameras. A collimation system must be designed to achieve the required resolution. Classical, geometric design is not satisfactory in the presence of scattering materials. For this reason a Monte Carlo simulation allowing flexible choice of collimator parameters and source distribution was developed. The simulation was fully tested and then used to predict the performance of collimators for probe and camera based systems. These assessments were carried out for the 'worst case source' which was a concept developed and validated to allow faster prediction of collimator performance. In essence the geometry of this source is such as to allow a resolution measurement to be made which represents the worst value expected from the system. The effect of changes in collimation on image quality was assessed using the computer phantoms and simulation of the data acquisition process on the singleslice system. These data were

  12. Complete photonic band gaps and tunable self-collimation in the two-dimensional plasma photonic crystals with a new structure

    SciTech Connect

    Zhang, Hai-Feng; Ding, Guo-Wen; Li, Hai-Ming; Liu, Shao-Bin

    2015-02-15

    In this paper, the properties of complete photonic band gaps (CPBGs) and tunable self-collimation in two-dimensional plasma photonic crystals (2D PPCs) with a new structure in square lattices, whose dielectric fillers (GaAs) are inserted into homogeneous and nomagnetized plasma background are theoretically investigated by a modified plane wave expansion (PWE) method with a novel technique. The novel PWE method can be utilized to compute the dispersion curves of 2D PPCs with arbitrary-shaped cross section in any lattices. As a comparison, CPBGs of PPCs for four different configurations are numerically calculated. The computed results show that the proposed design has the advantages of achieving the larger CPBGs compared to the other three configurations. The influences of geometric parameters of filled unit cell and plasma frequency on the properties of CPBGs are studied in detail. The calculated results demonstrate that CPBGs of the proposed 2D PPCs can be easily engineered by changing those parameters, and the larger CPBGs also can be obtained by optimization. The self-collimation in such 2D PPCs also is discussed in theory under TM wave. The theoretical simulations reveal that the self-collimation phenomena can be found in the TM bands, and both the frequency range of self-collimation and the equifrequency surface contours can be tuned by the parameters as mentioned above. It means that the frequency range and direction of electromagnetic wave can be manipulated by designing, as it propagates in the proposed PPCs without diffraction. Those results can hold promise for designing the tunable applications based on the proposed PPCs.

  13. Complete photonic band gaps and tunable self-collimation in the two-dimensional plasma photonic crystals with a new structure

    NASA Astrophysics Data System (ADS)

    Zhang, Hai-Feng; Ding, Guo-Wen; Li, Hai-Ming; Liu, Shao-Bin

    2015-02-01

    In this paper, the properties of complete photonic band gaps (CPBGs) and tunable self-collimation in two-dimensional plasma photonic crystals (2D PPCs) with a new structure in square lattices, whose dielectric fillers (GaAs) are inserted into homogeneous and nomagnetized plasma background are theoretically investigated by a modified plane wave expansion (PWE) method with a novel technique. The novel PWE method can be utilized to compute the dispersion curves of 2D PPCs with arbitrary-shaped cross section in any lattices. As a comparison, CPBGs of PPCs for four different configurations are numerically calculated. The computed results show that the proposed design has the advantages of achieving the larger CPBGs compared to the other three configurations. The influences of geometric parameters of filled unit cell and plasma frequency on the properties of CPBGs are studied in detail. The calculated results demonstrate that CPBGs of the proposed 2D PPCs can be easily engineered by changing those parameters, and the larger CPBGs also can be obtained by optimization. The self-collimation in such 2D PPCs also is discussed in theory under TM wave. The theoretical simulations reveal that the self-collimation phenomena can be found in the TM bands, and both the frequency range of self-collimation and the equifrequency surface contours can be tuned by the parameters as mentioned above. It means that the frequency range and direction of electromagnetic wave can be manipulated by designing, as it propagates in the proposed PPCs without diffraction. Those results can hold promise for designing the tunable applications based on the proposed PPCs.

  14. Un-collimated single-photon imaging system for high-sensitivity small animal and plant imaging

    NASA Astrophysics Data System (ADS)

    Walker, Katherine L.; Judenhofer, Martin S.; Cherry, Simon R.; Mitchell, Gregory S.

    2015-01-01

    In preclinical single-photon emission computed tomography (SPECT) system development the primary objective has been to improve spatial resolution by using novel parallel-hole or multi-pinhole collimator geometries. However, such high-resolution systems have relatively poor sensitivity (typically 0.01-0.1%). In contrast, a system that does not use collimators can achieve very high-sensitivity. Here we present a high-sensitivity un-collimated detector single-photon imaging (UCD-SPI) system for the imaging of both small animals and plants. This scanner consists of two thin, closely spaced, pixelated scintillator detectors that use NaI(Tl), CsI(Na), or BGO. The performance of the system has been characterized by measuring sensitivity, spatial resolution, linearity, detection limits, and uniformity. With 99mTc (140 keV) at the center of the field of view (20 mm scintillator separation), the sensitivity was measured to be 31.8% using the NaI(Tl) detectors and 40.2% with CsI(Na). The best spatial resolution (FWHM when the image formed as the geometric mean of the two detector heads, 20 mm scintillator separation) was 19.0 mm for NaI(Tl) and 11.9 mm for CsI(Na) at 140 keV, and 19.5 mm for BGO at 1116 keV, which is somewhat degraded compared to the cm-scale resolution obtained with only one detector head and a close source. The quantitative accuracy of the system’s linearity is better than 2% with detection down to activity levels of 100 nCi. Two in vivo animal studies (a renal scan using 99mTc MAG-3 and a thyroid scan with 123I) and one plant study (a 99mTcO4- xylem transport study) highlight the unique capabilities of this UCD-SPI system. From the renal scan, we observe approximately a one thousand-fold increase in sensitivity compared to the Siemens Inveon SPECT/CT scanner. UCD-SPI is useful for many imaging tasks that do not require excellent spatial resolution, such as high-throughput screening applications, simple radiotracer uptake studies in tumor

  15. Statistical performance evaluation and comparison of a Compton medical imaging system and a collimated Anger camera for higher energy photon imaging

    NASA Astrophysics Data System (ADS)

    Han, Li; Rogers, W. Leslie; Huh, Sam S.; Clinthorne, Neal

    2008-12-01

    In radionuclide treatment, tumor cells are primarily destroyed by charged particles emitted by the compound while associated higher energy photons are used to image the tumor in order to determine radiation dose and monitor shrinkage. However, the higher energy photons are difficult to image with conventional collimated Anger cameras, since a tradeoff exists between resolution and sensitivity, and the collimator septal penetration and scattering is increased due to the high energy photons. This research compares imaging performance of the conventional Anger camera to a Compton imaging system that can have improved spatial resolution and sensitivity for high energy photons because this tradeoff is decoupled, and the effect of Doppler broadening at higher gamma energies is decreased. System performance is analyzed by the modified uniform Cramer-Rao bound (M-UCRB) algorithms based on the developed system modeling. The bound shows that the effect of Doppler broadening is the limiting factor for Compton camera performance for imaging 364.4 keV photons emitted from 131I. According to the bound, the Compton camera outperforms the collimated system for an equal number of detected events when the desired spatial resolution for a 26 cm diameter uniform disk object is better than 12 mm FWHM. For a 3D cylindrical phantom, the lower bound on variance for the collimated camera is greater than for the Compton imaginer over the resolution range from 0.5 to 2 cm FWHM. Furthermore, the detection sensitivity of the proposed Compton imaging system is about 15-20 times higher than that of the collimated Anger camera.

  16. Different collimators in convergent beam irradiation of irregularly shaped intracranial target volumes.

    PubMed

    Otto-Oelschläger, S; Schlegel, W; Lorenz, W

    1994-02-01

    We compare different collimator forms (circular, elliptic and multi-leaf) in 3-D multiple arc rotation therapy for irregularly shaped intracranial tumors. When homogeneous irradiation of the tumor is ensured, the efficiency of treatment is expressed by the sparing of normal tissue outside the target volume to high dose irradiation. By utilizing integral dose-volume histograms we demonstrate that the multi-leaf collimator has considerable advantages. PMID:8184118

  17. Focusing, collimation and flux throughput at the IMCA-CAT bending-magnet beamline at the Advanced Photon Source.

    PubMed

    Koshelev, Irina; Huang, Rong; Graber, Timothy; Meron, Mati; Muir, J Lewis; Lavender, William; Battaile, Kevin; Mulichak, Anne M; Keefe, Lisa J

    2009-09-01

    The IMCA-CAT bending-magnet beamline was upgraded with a collimating mirror in order to achieve the energy resolution required to conduct high-quality multi- and single-wavelength anomalous diffraction (MAD/SAD) experiments without sacrificing beamline flux throughput. Following the upgrade, the bending-magnet beamline achieves a flux of 8 x 10(11) photons s(-1) at 1 A wavelength, at a beamline aperture of 1.5 mrad (horizontal) x 86 microrad (vertical), with energy resolution (limited mostly by the intrinsic resolution of the monochromator optics) deltaE/E = 1.5 x 10(-4) (at 10 kV). The beamline operates in a dynamic range of 7.5-17.5 keV and delivers to the sample focused beam of size (FWHM) 240 microm (horizontally) x 160 microm (vertically). The performance of the 17-BM beamline optics and its deviation from ideally shaped optics is evaluated in the context of the requirements imposed by the needs of protein crystallography experiments. An assessment of flux losses is given in relation to the (geometric) properties of major beamline components.

  18. Analysing collimator structure effects in head-scatter calculations for IMRT class fields using scatter raytracing.

    PubMed

    Naqvi, S A; Sarfaraz, M; Holmes, T; Yu, C X; Li, X A

    2001-07-01

    The frequent blocking of the irradiated volume in intensity modulated radiation therapy (IMRT) makes the head-scatter fraction of the incident photon fluence more significant than that in conventional therapy with open fields. On the other hand. certain collimator configurations block scatter photons directed to a given observation point while allowing primary photons to be transmitted. The 'anomalous blocking' makes the primary field a poor indicator of the scatter fluence. Since large MU-to-cGy ratios in IMRT can magnify head-scatter uncertainties, it becomes necessary to accurately model both the effective scatter source and the collimator structure that limits the scatter reaching the irradiated volume. First we obtain a dual-source model, using a Taylor series expansion to derive the effective scatter source distribution from the data measured for the Elekta SL20 linac equipped with a multi-leaf collimator (MLC). Then, using a raytracing algorithm, we calculate the transmission of scatter rays from the effective scatter source plane to points in the patient plane. The method can account for the anomalous blocking of scatter by the MLC leaves and the backup diaphragms. For a variety of collimator settings tested, the calculations agree with measurements to an accuracy of 0.002psi10 x 10, where psi10 x 10 is the total (primary + scatter) photon fluence of an open 10 x 10 cm2 field for the same MU delivered. Although the significance of collimator structure in IMRT depends strongly on fields shapes employed for the delivery, potential cumulative errors on the order of a few per cent can be avoided in fluence calculations if the proposed method is used. PMID:11474941

  19. Monte Carlo simulation of the photoneutron field in linac radiotherapy treatments with different collimation systems

    NASA Astrophysics Data System (ADS)

    Zanini, A.; Durisi, E.; Fasolo, F.; Ongaro, C.; Visca, L.; Nastasi, U.; Burn, K. W.; Scielzo, G.; Adler, J. O.; Annand, J. R. M.; Rosner, G.

    2004-02-01

    Bremsstrahlung photon beams produced by linac accelerators are currently the most commonly used method of radiotherapy for tumour treatments. When the photon energy exceeds 10 MeV the patient receives an undesired dose due to photoneutron production in the accelerator head. In the last few decades, new sophisticated techniques such as multileaf collimators have been used for a better definition of the target volume. In this case it is crucial to evaluate the photoneutron dose produced after giant dipole resonance (GDR) excitation of the high Z materials (mainly tungsten and lead) constituting the collimator leaves in view of the optimization of the radiotherapy treatment. A Monte Carlo approach has been used to calculate the photoneutron dose arising from the GDR reaction during radiotherapy with energetic photon beams. The simulation has been performed using the code MCNP4B-GN which is based on MCNP4B, but includes a new routine GAMMAN to model photoneutron production. Results for the facility at IRCC (Istituto per la Ricerca e la Cura del Cancro) Candiolo (Turin), which is based on 18 MV x-rays from a Varian Clinac 2300 C/D, are presented for a variety of different collimator configurations.

  20. Measurements of Output Factors For Small Photon Fields Up to 10 cm x 10 cm

    NASA Astrophysics Data System (ADS)

    Bacala, Angelina

    Field output factors (OF) for photon beams from a 6 MV medical accelerator were measured using five different detectors in a scanning water phantom. The measurements were taken for square field sizes of integral widths ranging from 1 cm to 10 cm for two reference source-to-surface distances (SSD) and depths in water. For the diode detectors, square field widths as small as 2.5 mm were also studied. The photon beams were collimated by using either the jaws or the multileaf collimators. Measured OFs are found to depend upon the field size, SSD, depth and also upon the type of beam collimation, size and type of detector used. For field sizes larger than 3 cm x 3 cm, the OF measurements agree to within 1% or less. The largest variation in OF occurs for jawsshaped field of size 1 cm x 1cm, where a difference of more than 18% is observed.

  1. SU-E-J-104: Single Photon Image From PET with Insertable SPECT Collimator for Boron Neutron Capture Therapy: A Feasibility Study

    SciTech Connect

    Jung, J; Yoon, D; Suh, T

    2014-06-01

    Purpose: The aim of our proposed system is to confirm the feasibility of extraction of two types of images from one positron emission tomography (PET) module with an insertable collimator for brain tumor treatment during the BNCT. Methods: Data from the PET module, neutron source, and collimator was entered in the Monte Carlo n-particle extended (MCNPX) source code. The coincidence events were first compiled on the PET detector, and then, the events of the prompt gamma ray were collected after neutron emission by using a single photon emission computed tomography (SPECT) collimator on the PET. The obtaining of full width at half maximum (FWHM) values from the energy spectrum was performed to collect effective events for reconstructed image. In order to evaluate the images easily, five boron regions in a brain phantom were used. The image profiles were extracted from the region of interest (ROI) of a phantom. The image was reconstructed using the ordered subsets expectation maximization (OSEM) reconstruction algorithm. The image profiles and the receiver operating characteristic (ROC) curve were compiled for quantitative analysis from the two kinds of reconstructed image. Results: The prompt gamma ray energy peak of 478 keV appeared in the energy spectrum with a FWHM of 41 keV (6.4%). On the basis of the ROC curve in Region A to Region E, the differences in the area under the curve (AUC) of the PET and SPECT images were found to be 10.2%, 11.7%, 8.2% (center, Region C), 12.6%, and 10.5%, respectively. Conclusion: We attempted to acquire the PET and SPECT images simultaneously using only PET without an additional isotope. Single photon images were acquired using an insertable collimator on a PET detector. This research was supported by the Leading Foreign Research Institute Recruitment Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science, Information and Communication Technologies (ICT) and Future Planning (MSIP)(Grant No

  2. Light propagation characteristics in photonic crystal fibers with α-power profiles of air hole diameter distributions and their application to fiber collimator

    NASA Astrophysics Data System (ADS)

    Yokota, Hirohisa; Higuchi, Keiichi; Imai, Yoh

    2016-08-01

    Light propagation characteristics in photonic crystal fibers (PCFs) with α-power profiles of air hole diameter distributions were theoretically investigated. It was clarified that the intensity peak of the beam propagating in the PCF with Gaussian beam excitation varied periodically with little power attenuation. It was found that the envelope of the periodic intensity variation depended on α. We theoretically demonstrated that the PCF with the α-power profile of the air hole diameter distribution could be applied to a collimator for a conventional PCF with uniform air holes in Gaussian beam excitation to reduce coupling loss, where a PCF of appropriate length with the α-power air hole diameter distribution was spliced to a conventional PCF. It was also found that the coupling efficiency was higher for a larger α.

  3. Radiation leakage dose from Elekta electron collimation system.

    PubMed

    Pitcher, Garrett M; Hogstrom, Kenneth R; Carver, Robert L

    2016-09-08

    This study provided baseline data required for a greater project, whose objective was to design a new Elekta electron collimation system having significantly lighter electron applicators with equally low out-of field leakage dose. Specifically, off-axis dose profiles for the electron collimation system of our uniquely configured Elekta Infinity accelerator with the MLCi2 treatment head were measured and calculated for two primary purposes: 1) to evaluate and document the out-of-field leakage dose in the patient plane and 2) to validate the dose distributions calculated using a BEAMnrc Monte Carlo (MC) model for out-of-field dose profiles. Off-axis dose profiles were measured in a water phantom at 100 cm SSD for 1 and 2 cm depths along the in-plane, cross-plane, and both diagonal axes using a cylindrical ionization chamber with the 10 × 10 and 20 × 20 cm2 applicators and 7, 13, and 20 MeV beams. Dose distributions were calculated using a previously developed BEAMnrc MC model of the Elekta Infinity accelerator for the same beam energies and applicator sizes and compared with measurements. Measured results showed that the in-field beam flatness met our acceptance criteria (± 3% on major and ±4% on diagonal axes) and that out-of-field mean and maximum percent leakage doses in the patient plane met acceptance criteria as specified by the International Electrotechnical Commission (IEC). Cross-plane out-of-field dose profiles showed greater leakage dose than in-plane profiles, attributed to the curved edges of the upper X-ray jaws and multileaf collimator. Mean leakage doses increased with beam energy, being 0.93% and 0.85% of maximum central axis dose for the 10 × 10 and 20 × 20 cm2 applicators, respectively, at 20 MeV. MC calculations predicted the measured dose to within 0.1% in most profiles outside the radiation field; however, excluding model-ing of nontrimmer applicator components led to calculations exceeding measured data by as much as 0.2% for some regions

  4. Radiation leakage dose from Elekta electron collimation system.

    PubMed

    Pitcher, Garrett M; Hogstrom, Kenneth R; Carver, Robert L

    2016-01-01

    This study provided baseline data required for a greater project, whose objective was to design a new Elekta electron collimation system having significantly lighter electron applicators with equally low out-of field leakage dose. Specifically, off-axis dose profiles for the electron collimation system of our uniquely configured Elekta Infinity accelerator with the MLCi2 treatment head were measured and calculated for two primary purposes: 1) to evaluate and document the out-of-field leakage dose in the patient plane and 2) to validate the dose distributions calculated using a BEAMnrc Monte Carlo (MC) model for out-of-field dose profiles. Off-axis dose profiles were measured in a water phantom at 100 cm SSD for 1 and 2 cm depths along the in-plane, cross-plane, and both diagonal axes using a cylindrical ionization chamber with the 10 × 10 and 20 × 20 cm2 applicators and 7, 13, and 20 MeV beams. Dose distributions were calculated using a previously developed BEAMnrc MC model of the Elekta Infinity accelerator for the same beam energies and applicator sizes and compared with measurements. Measured results showed that the in-field beam flatness met our acceptance criteria (± 3% on major and ±4% on diagonal axes) and that out-of-field mean and maximum percent leakage doses in the patient plane met acceptance criteria as specified by the International Electrotechnical Commission (IEC). Cross-plane out-of-field dose profiles showed greater leakage dose than in-plane profiles, attributed to the curved edges of the upper X-ray jaws and multileaf collimator. Mean leakage doses increased with beam energy, being 0.93% and 0.85% of maximum central axis dose for the 10 × 10 and 20 × 20 cm2 applicators, respectively, at 20 MeV. MC calculations predicted the measured dose to within 0.1% in most profiles outside the radiation field; however, excluding model-ing of nontrimmer applicator components led to calculations exceeding measured data by as much as 0.2% for some regions

  5. Patient-dependent beam-modifier physics in Monte Carlo photon dose calculations.

    PubMed

    Schach von Wittenau, A E; Bergstrom, P M; Cox, L J

    2000-05-01

    Model pencil-beam on slab calculations are used as well as a series of detailed calculations of photon and electron output from commercial accelerators to quantify level(s) of physics required for the Monte Carlo transport of photons and electrons in treatment-dependent beam modifiers, such as jaws, wedges, blocks, and multileaf collimators, in photon teletherapy dose calculations. The physics approximations investigated comprise (1) not tracking particles below a given kinetic energy, (2) continuing to track particles, but performing simplified collision physics, particularly in handling secondary particle production, and (3) not tracking particles in specific spatial regions. Figures-of-merit needed to estimate the effects of these approximations are developed, and these estimates are compared with full-physics Monte Carlo calculations of the contribution of the collimating jaws to the on-axis depth-dose curve in a water phantom. These figures of merit are next used to evaluate various approximations used in coupled photon/electron physics in beam modifiers. Approximations for tracking electrons in air are then evaluated. It is found that knowledge of the materials used for beam modifiers, of the energies of the photon beams used, as well as of the length scales typically found in photon teletherapy plans, allows a number of simplifying approximations to be made in the Monte Carlo transport of secondary particles from the accelerator head and beam modifiers to the isocenter plane.

  6. The use of intensity-modulated radiation therapy photon beams for improving the dose uniformity of electron beams shaped with MLC

    SciTech Connect

    Mosalaei, Homeira; Karnas, Scott; Shah, Sheel; Van Doodewaard, Sharon; Foster, Tim; Chen, Jeff

    2012-04-01

    Electrons are ideal for treating shallow tumors and sparing adjacent normal tissue. Conventionally, electron beams are collimated by cut-outs that are time-consuming to make and difficult to adapt to tumor shape throughout the course of treatment. We propose that electron cut-outs can be replaced using photon multileaf collimator (MLC). Two major problems of this approach are that the scattering of electrons causes penumbra widening because of a large air gap, and available commercial treatment planning systems (TPSs) do not support MLC-collimated electron beams. In this study, these difficulties were overcome by (1) modeling electron beams collimated by photon MLC for a commercial TPS, and (2) developing a technique to reduce electron beam penumbra by adding low-energy intensity-modulated radiation therapy (IMRT) photons (4 MV). We used blocks to simulate MLC shielding in the TPS. Inverse planning was used to optimize boost photon beams. This technique was applied to a parotid and a central nervous system (CNS) clinical case. Combined photon and electron plans were compared with conventional plans and verified using ion chamber, film, and a 2D diode array. Our studies showed that the beam penumbra for mixed beams with 90 cm source to surface distance (SSD) is comparable with electron applicators and cut-outs at 100 cm SSD. Our mixed-beam technique yielded more uniform dose to the planning target volume and lower doses to various organs at risk for both parotid and CNS clinical cases. The plans were verified with measurements, with more than 95% points passing the gamma criteria of 5% in dose difference and 5 mm for distance to agreement. In conclusion, the study has demonstrated the feasibility and potential advantage of using photon MLC to collimate electron beams with boost photon IMRT fields.

  7. Effect of the collimator angle on dosimetric verification of volumetric modulated arc therapy

    NASA Astrophysics Data System (ADS)

    Kim, Yong Ho; Park, Ha Ryung; Kim, Won Taek; Kim, Dong Won; Ki, Yongkan; Lee, Juhye; Bae, Jinsuk; Park, Dahl; Jeon, Hosang; Nam, Ji Ho

    2015-07-01

    The collimator is usually rotated when planning volumetric modulated arc therapy (VMAT) due to the leakage of radiation between the multi-leaf collimator (MLC) leaves. We studied the effect of the collimator angle on the results of dosimetric verification of VMAT plans for head and neck patients. We studied VMAT plans for 10 head and neck patients. We made two sets of VMAT plans for each patient. Each set was composed of 10 plans with collimator angles of 0, 5, 10, 15, 20, 25, 30, 35, 40, 45 degrees. Plans in the first set were optimized individually, and plans in the second set shared the 30-degree collimator-angle optimization. The two sets of plans were verified by using the 2-dimensional ion chamber array MatriXX (IBA Dosimetry, Germany). The comparisons between the calculation and the measurements were made by using a γ-index analysis. The γ-index (2%/2 mm) and (3%/3 mm) passing rates had negative correlations with the collimator angle. The maximum difference between γ-index (3%/3 mm) passing rates of different collimator angles for each patient ranged from 1.46% to 5.60% with an average of 3.67%. There were significant differences (maximum 5.6%) in the passing rates for different collimator angles. The results suggested that the accuracy of the delivered dose depended on the collimator angle. These findings are informative when choosing a collimator angle for VMAT plans.

  8. Hybrid collimation for industrial gamma-ray imaging.

    SciTech Connect

    He, Z.; Knoll, G. F.; Smith, L. E.; Wehe, D. K.

    1999-06-22

    Portable photon imaging devices with a broad energy range of sensitivity, adequate angular resolution and high efficiency are useful in applications such as environmental remediation and industrial surveys. The vast majority of past systems built for these applications have relied on mechanical collimation although a few have used electronic collimation. To our knowledge, no devices have been built that exploit the benefits of both mechanical and electronic collimation in the same system. The combination of a mechanically-collimated camera with an electronically-collimated camera offers both the high efficiency and good angular resolution typical in a mechanically-collimated camera for lower energies and the uncoupling of spatial resolution and efficiency provided by an electronically-collimated camera at higher energies.

  9. Retinoblastoma external beam photon irradiation with a special ‘D’-shaped collimator: a comparison between measurements, Monte Carlo simulation and a treatment planning system calculation

    NASA Astrophysics Data System (ADS)

    Brualla, L.; Mayorga, P. A.; Flühs, A.; Lallena, A. M.; Sempau, J.; Sauerwein, W.

    2012-11-01

    Retinoblastoma is the most common eye tumour in childhood. According to the available long-term data, the best outcome regarding tumour control and visual function has been reached by external beam radiotherapy. The benefits of the treatment are, however, jeopardized by a high incidence of radiation-induced secondary malignancies and the fact that irradiated bones grow asymmetrically. In order to better exploit the advantages of external beam radiotherapy, it is necessary to improve current techniques by reducing the irradiated volume and minimizing the dose to the facial bones. To this end, dose measurements and simulated data in a water phantom are essential. A Varian Clinac 2100 C/D operating at 6 MV is used in conjunction with a dedicated collimator for the retinoblastoma treatment. This collimator conforms a ‘D’-shaped off-axis field whose irradiated area can be either 5.2 or 3.1 cm2. Depth dose distributions and lateral profiles were experimentally measured. Experimental results were compared with Monte Carlo simulations’ run with the penelope code and with calculations performed with the analytical anisotropic algorithm implemented in the Eclipse treatment planning system using the gamma test. penelope simulations agree reasonably well with the experimental data with discrepancies in the dose profiles less than 3 mm of distance to agreement and 3% of dose. Discrepancies between the results found with the analytical anisotropic algorithm and the experimental data reach 3 mm and 6%. Although the discrepancies between the results obtained with the analytical anisotropic algorithm and the experimental data are notable, it is possible to consider this algorithm for routine treatment planning of retinoblastoma patients, provided the limitations of the algorithm are known and taken into account by the medical physicist and the clinician. Monte Carlo simulation is essential for knowing these limitations. Monte Carlo simulation is required for optimizing the

  10. Characterisation of fan-beam collimators.

    PubMed

    Pareto, D; Pavía, J; Falcón, C; Juvells, I; Cot, A; Ros, D

    2001-02-01

    Fan-beam collimators offer a good balance between resolution and noise. The collimator response may be included in iterative reconstruction algorithms in order to improve single-photon emission tomography (SPET) resolution. To this end, accurate determination of the focal region and characterisation of the collimator response as a function of the source co-ordinates must be performed. In this paper, a method to characterise fanbeam collimators is evaluated. First, we calculated the real focal region and the accuracy of the collimator convergence. Then, we confirmed the hypothesis that Gaussian distributions adequately fit the collimator responses, although no individualised treatment was performed for the tails of detector response which are associated with scattering and septal penetration. Finally, analytical functions were used to model the resolution and sensitivity. The parameter values in these functions were obtained from experimental measures by non-linear regression fitting. Our findings show differences of 1.43% between nominal and real focal length and standard deviations of 2.5 mm in the x-direction and 7.1 mm in the y-direction for the focal convergence. The correlation coefficients between experimental and predicted values were 0.994 for resolution and 0.991 for sensitivity. As a consequence, the proposed method can be used to characterise the collimator response.

  11. Conformity of LINAC-Based Stereotactic Radiosurgery Using Dynamic Conformal Arcs and Micro-Multileaf Collimator

    SciTech Connect

    Hazard, Lisa J. Wang, Brian; Skidmore, Thomas B.; Chern, Shyh-Shi; Salter, Bill J.; Jensen, Randy L.; Shrieve, Dennis C.

    2009-02-01

    Purpose: To assess the conformity of dynamic conformal arc linear accelerator-based stereotactic radiosurgery and to describe a standardized method of isodose surface (IDS) selection. Methods and Materials: In 174 targets, the conformity index (CI) at the prescription IDS used for treatment was calculated as CI = (PIV/PVTV)/(PVTV/TV), where TV is the target volume, PIV (prescription isodose volume) is the total volume encompassed by the prescription IDS, and PVTV is the TV encompassed by the IDS. In addition, a 'standardized' prescription IDS (sIDS) was chosen according to the following criteria: 95% of the TV was encompassed by the PIV and 99% of TV was covered by 95% of the prescription dose. The CIs at the sIDS were also calculated. Results: The median CI at the prescription IDS and sIDS was 1.63 and 1.47, respectively (p < 0.001). In 132 of 174 cases, the volume of normal tissue in the PIV was reduced by the prescription to the sIDS compared with the prescription IDS, in 20 cases it remained unchanged, and in 22 cases it was increased. Conclusion: The CIs obtained with linear accelerator-based stereotactic radiosurgery are comparable to those previously reported for gamma knife stereotactic radiosurgery. Using a uniform method to select the sIDS, adequate target coverage was usually achievable with prescription to an IDS greater than that chosen by the treating physician (prescription IDS), providing sparing of normal tissue. Thus, the sIDS might aid physicians in identifying a prescription IDS that balances coverage and conformity.

  12. Radiation collimator and systems incorporating same

    DOEpatents

    Norman, Daren R.; Yoon, Woo Y.; Jones, James L.; Haskell, Kevin J.; Bennett, Brion D.; Tschaggeny, Charles W.; Jones, Warren F.

    2011-09-13

    A collimator including a housing having disposed therein a shield element surrounding a converter core in which a photon beam is generated from electrons emanating from a linear accelerator. A beam channeler longitudinally adjacent the shield element has a beam aperture therethrough coaxially aligned with, and of the same diameter as, an exit bore of the converter core. A larger entry bore in the converter core is coaxial with, and longitudinally separated from, the exit bore thereof. Systems incorporating the collimator are also disclosed.

  13. An energy-optimized collimator design for a CZT-based SPECT camera

    PubMed Central

    Weng, Fenghua; Bagchi, Srijeeta; Zan, Yunlong; Huang, Qiu; Seo, Youngho

    2015-01-01

    In single photon emission computed tomography, it is a challenging task to maintain reasonable performance using only one specific collimator for radio-tracers over a broad spectrum of diagnostic photon energies, since photon scatter and penetration in a collimator differ with the photon energy. Frequent collimator exchanges are inevitable in daily clinical SPECT imaging, which hinders throughput while subjecting the camera to operational errors and damage. Our objective is to design a collimator, which independent of the photon energy performs reasonably well for commonly used radiotracers with low- to medium-energy levels of gamma emissions. Using the Geant4 simulation toolkit, we simulated and evaluated a parallel-hole collimator mounted to a CZT detector. With the pixel-geometry-matching collimation, the pitch of the collimator hole was fixed to match the pixel size of the CZT detector throughout this work. Four variables, hole shape, hole length, hole radius/width and the source-to-collimator distance were carefully studied. Scatter and penetration of the collimator, sensitivity and spatial resolution of the system were assessed for four radionuclides including 57Co, 99mTc, 123I and 111In, with respect to the aforementioned four variables. An optimal collimator was then decided upon such that it maximized the total relative sensitivity (TRS) for the four considered radionuclides while other performance parameters, such as scatter, penetration and spatial resolution, were benchmarked to prevalent commercial scanners and collimators. Digital phantom studies were also performed to validate the system with the optimal square-hole collimator (23 mm hole length, 1.28 mm hole width, 0.32 mm septal thickness) in terms of contrast, contrast-to-noise ratio and recovery ratio. This study demonstrates promise of our proposed energy-optimized collimator to be used in a CZT-based gamma camera, with comparable or even better imaging performance versus commercial collimators

  14. Maximizing the potential of direct aperture optimization through collimator rotation

    SciTech Connect

    Milette, Marie-Pierre; Otto, Karl

    2007-04-15

    Intensity-modulated radiation therapy (IMRT) treatment plans are conventionally produced by the optimization of fluence maps followed by a leaf sequencing step. An alternative to fluence based inverse planning is to optimize directly the leaf positions and field weights of multileaf collimator (MLC) apertures. This approach is typically referred to as direct aperture optimization (DAO). It has been shown that equivalent dose distributions may be generated that have substantially fewer monitor units (MU) and number of apertures compared to fluence based optimization techniques. Here we introduce a DAO technique with rotated apertures that we call rotating aperture optimization (RAO). The advantages of collimator rotation in IMRT have been shown previously and include higher fluence spatial resolution, increased flexibility in the generation of aperture shapes and less interleaf effects. We have tested our RAO algorithm on a complex C-shaped target, seven nasopharynx cancer recurrences, and one multitarget nasopharynx carcinoma patient. A study was performed in order to assess the capabilities of RAO as compared to fixed collimator angle DAO. The accuracy of fixed and rotated collimator aperture delivery was also verified. An analysis of the optimized treatment plans indicates that plans generated with RAO are as good as or better than DAO while maintaining a smaller number of apertures and MU than fluence based IMRT. Delivery verification results show that RAO is less sensitive to tongue and groove effects than DAO. Delivery time is currently increased due to the collimator rotation speed although this is a mechanical limitation that can be eliminated in the future.

  15. Optimization of convergent collimators for pixelated SPECT systems

    SciTech Connect

    Capote, Ricardo M.; Matela, Nuno; Conceicao, Raquel C.; Almeida, Pedro

    2013-06-15

    Purpose: The optimization of the collimator design is essential to obtain the best possible sensitivity in single photon emission computed tomography imaging. The aim of this work is to present a methodology for maximizing the sensitivity of convergent collimators, specifically designed to match the pitch of pixelated detectors, for a fixed spatial resolution value and to present some initial results using this approach. Methods: Given the matched constraint, the optimal collimator design cannot be simply found by allowing the highest level of septal penetration and spatial resolution consistent with the imposed restrictions, as it is done for the optimization of conventional collimators. Therefore, an algorithm that interactively calculates the collimator dimensions, with the maximum sensitivity, which respect the imposed restrictions was developed and used to optimize cone and fan beam collimators with tapered square-shaped holes for low (60-300 keV) and high energy radiation (300-511 keV). The optimal collimator dimensions were locally calculated based on the premise that each hole and septa of the convergent collimator should locally resemble an appropriate optimal matched parallel collimator. Results: The optimal collimator dimensions, calculated for subcentimeter resolutions (3 and 7.5 mm), common pixel sizes (1.6, 2.1, and 2.5 mm), and acceptable septal penetration at 140 keV, were approximately constant throughout the collimator, despite their different hole incidence angles. By using these input parameters and a less strict septal penetration value of 5%, the optimal collimator dimensions and the corresponding mass per detector area were calculated for 511 keV. It is shown that a low value of focal distance leads to improvements in the average sensitivity at a fixed source-collimator distance and resolution. The optimal cone beam performance outperformed that of other optimal collimation geometries (fan and parallel beam) in imaging objects close to the

  16. COLLIMATION EXPERIENCE AT RHIC.

    SciTech Connect

    DREES,K.A.FLILLER,R.TRBOJEVIC,D.KAIN,V.

    2003-05-19

    In the Relativistic Heavy Ion Collider (RHIC) the abort kicker magnets are the limiting aperture. Continuous losses at this location could deteriorate the kicker performance. In addition, losses especially in the triplet area cause backgrounds in the experimental detectors. The RHIC one-stage collimation system was used to reduce these backgrounds as well as losses at the abort kickers. Collimation performance and results from various runs with even and uneven species (Au-Au, pp and d-Au) are presented and compared. Upgrades of the system for the upcoming high luminosity runs are outlined.

  17. Determination of optimal collimation parameters for a rotating slat collimator system: a system matrix method using ML-EM

    NASA Astrophysics Data System (ADS)

    Boisson, F.; Bekaert, V.; Brasse, D.

    2016-03-01

    Nowadays, Single Photon imaging has become an essential part of molecular imaging and nuclear medicine. Whether to establish a diagnosis or in the therapeutic monitoring, this modality presents performance that continues to improve. For over 50 years, several collimators have been proposed. Mainly governed by collimation parameters, the resolution-sensitivity trade-off is the factor determining the collimator the most suitable for an intended study. One alternative to the common approaches is the rotating slat collimator (RSC). In the present study, we are aiming at developing a preclinical system equipped with a RSC dedicated to mice and rats imaging, which requires both high sensitivity and spatial resolution. We investigated the resolution-sensitivity trade-offs obtained by varying different collimation parameters: (i) the slats height (H), and (ii) the gap between two consecutive slats (g), considering different intrinsic spatial resolutions. One system matrix was generated for each set of collimation parameters (H,g). Spatial resolutions, Signal-to-Noise Ratio (SNR) and sensitivity obtained for all the set of collimation parameters (H,g) were measured in the 2D projections reconstructed with ML-EM. According to our results, 20 mm high slats and a 1 mm gap were chosen as a good RSC candidate for a preclinical detection module. This collimator will ensure a sensitivity greater than 0.2% and a system spatial resolution below 1 mm, considering an intrinsic spatial resolution below 0.8 mm.

  18. Determination of optimal collimation parameters for a rotating slat collimator system: a system matrix method using ML-EM.

    PubMed

    Boisson, F; Bekaert, V; Brasse, D

    2016-03-21

    Nowadays, Single Photon imaging has become an essential part of molecular imaging and nuclear medicine. Whether to establish a diagnosis or in the therapeutic monitoring, this modality presents performance that continues to improve. For over 50 years, several collimators have been proposed. Mainly governed by collimation parameters, the resolution-sensitivity trade-off is the factor determining the collimator the most suitable for an intended study. One alternative to the common approaches is the rotating slat collimator (RSC). In the present study, we are aiming at developing a preclinical system equipped with a RSC dedicated to mice and rats imaging, which requires both high sensitivity and spatial resolution. We investigated the resolution-sensitivity trade-offs obtained by varying different collimation parameters: (i) the slats height (H), and (ii) the gap between two consecutive slats (g), considering different intrinsic spatial resolutions. One system matrix was generated for each set of collimation parameters (H,g). Spatial resolutions, Signal-to-Noise Ratio (SNR) and sensitivity obtained for all the set of collimation parameters (H,g) were measured in the 2D projections reconstructed with ML-EM. According to our results, 20 mm high slats and a 1 mm gap were chosen as a good RSC candidate for a preclinical detection module. This collimator will ensure a sensitivity greater than 0.2% and a system spatial resolution below 1 mm, considering an intrinsic spatial resolution below 0.8 mm. PMID:26930449

  19. Multipinhole collimator with 20 apertures for a brain SPECT application

    SciTech Connect

    Lee, Tzu-Cheng; Ellin, Justin R.; Shrestha, Uttam; Seo, Youngho; Huang, Qiu; Gullberg, Grant T.

    2014-11-01

    Purpose: Several new technologies for single photon emission computed tomography (SPECT) instrumentation with parallel-hole collimation have been proposed to improve detector sensitivity and signal collection efficiency. Benefits from improved signal efficiency include shorter acquisition times and lower dose requirements. In this paper, the authors show a possibility of over an order of magnitude enhancement in photon detection efficiency (from 7.6 × 10{sup −5} to 1.6 × 10{sup −3}) for dopamine transporter (DaT) imaging of the striatum over the conventional SPECT parallel-hole collimators by use of custom-designed 20 multipinhole (20-MPH) collimators with apertures of 0.75 cm diameter. Methods: Quantifying specific binding ratio (SBR) of {sup 123}I-ioflupane or {sup 123}I-iometopane’s signal at the striatal region is a common brain imaging method to confirm the diagnosis of the Parkinson’s disease. The authors performed imaging of a striatal phantom filled with aqueous solution of I-123 and compared camera recovery ratios of SBR acquired between low-energy high-resolution (LEHR) parallel-hole collimators and 20-MPH collimators. Results: With only two-thirds of total acquisition time (20 min against 30 min), a comparable camera recovery ratio of SBR was achieved using 20-MPH collimators in comparison to that from the LEHR collimator study. Conclusions: Their systematic analyses showed that the 20-MPH collimator could be a promising alternative for the DaT SPECT imaging for brain over the traditional LEHR collimator, which could give both shorter scan time and improved diagnostic accuracy.

  20. Enhanced PET resolution by combining pinhole collimation and coincidence detection.

    PubMed

    DiFilippo, Frank P

    2015-10-21

    Spatial resolution of clinical PET scanners is limited by detector design and photon non-colinearity. Although dedicated small animal PET scanners using specialized high-resolution detectors have been developed, enhancing the spatial resolution of clinical PET scanners is of interest as a more available alternative. Multi-pinhole 511 keV SPECT is capable of high spatial resolution but requires heavily shielded collimators to avoid significant background counts. A practical approach with clinical PET detectors is to combine multi-pinhole collimation with coincidence detection. In this new hybrid modality, there are three locations associated with each event, namely those of the two detected photons and the pinhole aperture. These three locations over-determine the line of response and provide redundant information that is superior to coincidence detection or pinhole collimation alone. Multi-pinhole collimation provides high resolution and avoids non-colinearity error but is subject to collimator penetration and artifacts from overlapping projections. However the coincidence information, though at lower resolution, is valuable for determining whether the photon passed near a pinhole within the cone acceptance angle and for identifying through which pinhole the photon passed. This information allows most photons penetrating through the collimator to be rejected and avoids overlapping projections. With much improved event rejection, a collimator with minimal shielding may be used, and a lightweight add-on collimator for high resolution imaging is feasible for use with a clinical PET scanner. Monte Carlo simulations were performed of a (18)F hot rods phantom and a 54-pinhole unfocused whole-body mouse collimator with a clinical PET scanner. Based on coincidence information and pinhole geometry, events were accepted or rejected, and pinhole-specific crystal-map projections were generated. Tomographic images then were reconstructed using a conventional pinhole SPECT

  1. Optimization of Collimator Trajectory in Volumetric Modulated Arc Therapy: Development and Evaluation for Paraspinal SBRT

    SciTech Connect

    Zhang Pengpeng; Happersett, Laura; Yang Yingli; Yamada, Yoshiya; Mageras, Gig; Hunt, Margie

    2010-06-01

    Purpose: To develop a collimator trajectory optimization paradigm for volumetric modulated arc therapy (VMAT) and evaluate this technique in paraspinal stereotactic body radiation therapy (SBRT). Method and Materials: We propose a novel VMAT paradigm, Coll-VMAT, which integrates collimator rotation with synchronized gantry rotation, multileaf collimator (MLC) motion, and dose-rate modulation. At each gantry angle a principal component analysis (PCA) is applied to calculate the primary cord orientation. The collimator angle is then aligned so that MLC travel is parallel to the PCA-derived direction. An in-house VMAT optimization follows the geometry-based collimator trajectory optimization to obtain the optimal MLC position and monitor units (MU) at each gantry angle. A treatment planning study of five paraspinal SBRT patients compared Coll-VMAT to standard VMAT (fixed collimator angle) and static field IMRT plans. Plan evaluation statistics included planning target volume (PTV) V95%, PTV-D95%, cord-D05%, and total beam-on time. Results: Variation of collimator angle in Coll-VMAT plans ranges from 26 deg. to 54 deg., with a median of 40 deg. Patient-averaged PTV V95% (94.6% Coll-VMAT vs. 92.1% VMAT and 93.3% IMRT) and D95% (22.5 Gy vs. 21.4 Gy and 22.0 Gy, respectively) are highest with Coll-VMAT, and cord D05% (9.8 Gy vs. 10.0 Gy and 11.7 Gy) is lowest. Total beam-on time with Coll-VMAT (5,164 MU) is comparable to standard VMAT (4,868 MU) and substantially lower than IMRT (13,283 MU). Conclusion: Collimator trajectory optimization-based VMAT provides an additional degree of freedom that can improve target coverage and cord sparing of paraspinal SBRT plans compared with standard VMAT and IMRT approaches.

  2. VIRUS instrument collimator assembly

    NASA Astrophysics Data System (ADS)

    Marshall, Jennifer L.; DePoy, Darren L.; Prochaska, Travis; Allen, Richard D.; Williams, Patrick; Rheault, Jean-Philippe; Li, Ting; Nagasawa, Daniel Q.; Akers, Christopher; Baker, David; Boster, Emily; Campbell, Caitlin; Cook, Erika; Elder, Alison; Gary, Alex; Glover, Joseph; James, Michael; Martin, Emily; Meador, Will; Mondrik, Nicholas; Rodriguez-Patino, Marisela; Villanueva, Steven; Hill, Gary J.; Tuttle, Sarah; Vattiat, Brian; Lee, Hanshin; Chonis, Taylor S.; Dalton, Gavin B.; Tacon, Mike

    2014-07-01

    The Visual Integral-Field Replicable Unit Spectrograph (VIRUS) instrument is a baseline array 150 identical fiber fed optical spectrographs designed to support observations for the Hobby-Eberly Telescope Dark Energy Experiment (HETDEX). The collimator subassemblies of the instrument have been assembled in a production line and are now complete. Here we review the design choices and assembly practices used to produce a suite of identical low-cost spectrographs in a timely fashion using primarily unskilled labor.

  3. Gamma ray collimator

    NASA Technical Reports Server (NTRS)

    Casanova, Edgar J. (Inventor)

    1991-01-01

    A gamma ray collimator including a housing having first and second sections is disclosed. The first section encloses a first section of depleted uranium which is disposed for receiving and supporting a radiation emitting component such as cobalt 60. The second section encloses a depleted uranium member which is provided with a conical cut out focusing portion disposed in communication with the radiation emitting element for focusing the emitted radiation to the target.

  4. Choreographing Couch and Collimator in Volumetric Modulated Arc Therapy

    SciTech Connect

    Yang Yingli; Zhang Pengpeng; Happersett, Laura; Xiong Jianping; Yang Jie; Chan, Maria; Beal, Kathryn; Mageras, Gig; Hunt, Margie

    2011-07-15

    Purpose: To design and optimize trajectory-based, noncoplanar subarcs for volumetric modulated arc therapy (VMAT) deliverable on both Varian TrueBEAM system and traditional accelerators; and to investigate their potential advantages for treating central nervous system (CNS) tumors. Methods and Materials: To guide the computerized selection of beam trajectories consisting of simultaneous couch, gantry, and collimator motion, a score function was implemented to estimate the geometric overlap between targets and organs at risk for each couch/gantry angle combination. An initial set of beam orientations is obtained as a function of couch and gantry angle, according to a minimum search of the score function excluding zones of collision. This set is grouped into multiple continuous and extended subarcs subject to mechanical limitations using a hierarchical clustering algorithm. After determination of couch/gantry trajectories, a principal component analysis finds the collimator angle at each beam orientation that minimizes residual target-organ at risk overlaps. An in-house VMAT optimization algorithm determines the optimal multileaf collimator position and monitor units for control points within each subarc. A retrospective study of 10 CNS patients compares the proposed method of VMAT trajectory with dynamic gantry, leaves, couch, and collimator motion (Tra-VMAT); a standard noncoplanar VMAT with no couch/collimator motion within subarcs (Std-VMAT); and noncoplanar intensity-modulated radiotherapy (IMRT) plans that were clinically used. Results: Tra-VMAT provided improved target dose conformality and lowered maximum dose to brainstem, optic nerves, and chiasm by 7.7%, 1.1%, 2.3%, and 1.7%, respectively, compared with Std-VMAT. Tra-VMAT provided higher planning target volume minimum dose and reduced maximum dose to chiasm, optic nerves, and cochlea by 6.2%, 1.3%, 6.3%, and 8.4%, respectively, and reduced cochlea mean dose by 8.7%, compared with IMRT. Tra-VMAT averaged

  5. Commissioning of mini-multi-leaf-collimator (MMLC) for stereotactic radiosurgery and radiotherapy.

    PubMed

    Mardirossian, George; Urie, Marcia; Fitzgerald, Thomas J; Mayr, Nina; Montebello, Joseph; Lo, Yuan-Chynan

    2003-01-01

    Commissioning of a Radionics miniature multi-leaf collimator (MMLC) for stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT) is reported. With single isocenter and multi static fields, the MMLC can provide better conformity of dose distributions to the target and/or irregularly shaped target volumes than standard arc (circular) field beams with multiple isocenters. Advantages offered by the MMLC over traditional LINAC based SRS and SRT includes greatly improved dose homogeneity to the target, reduced patient positioning time and reduced treatment time. In this work, the MMLC is attached to a Varian 2300 C/D with Varian 80-leaf multi-leaf collimator. The MMLC has 62 leaves, each measured to a width of 3.53 mm at isocenter, with fields range from 1x1 cm to less than 10 × 12 cm. Beam parameters required by the Radionics treatment planning system (XPlan version 2) for evaluating the dose include tissue maximum ratio (TMR), scatter factors (SF), off-axis ratios (OAR), output factors, penumbra function (P) and transmission factors (TF) are performed in this work. Beam data are acquired with a small stereotactic diode, standard ion chambers and radiographic films. Measured profiles of dose distribution are compared to those calculated by the software and absolute dosimetry is performed. PMID:22388095

  6. Experimentally simulating quantum walks with self-collimated light

    PubMed Central

    Qi, F.; Wang, Y. F.; Ma, Q. Y.; Zheng, W. H.

    2016-01-01

    In self-collimated photonic crystal, periodically arranged air holes of sub-wavelength scale provide flattened equi-frequency curves perpendicular to the ΓM direction, which allow light or photons propagating in a quasi-uniform medium without diffraction. Here we for the first time experimentally simulate four-step single-photon discrete time quantum walks with classical light in such a photonic crystal chip fabricated on silicon-on-insulator. Similarities between theoretical expectations and experimental results are higher than 0.98. The functional area is compact and can be extended to construct more complicated linear quantum circuits. PMID:27353428

  7. A dosimetric characterization of a novel linear accelerator collimator

    SciTech Connect

    Thompson, C. M.; Weston, S. J. Cosgrove, V. C.; Thwaites, D. I.

    2014-03-15

    Purpose: The aim of this work is to characterize a new linear accelerator collimator which contains a single pair of sculpted diaphragms mounted orthogonally to a 160 leaf multileaf collimator (MLC). The diaphragms have “thick” regions providing full attenuation and “thin” regions where attenuation is provided by both the leaves and the diaphragm. The leaves are mounted on a dynamic leaf guide allowing rapid leaf motion and leaf travel over 350 mm. Methods: Dosimetric characterization, including assessment of leaf transmission, leaf tip transmission, penumbral width, was performed in a plotting tank. Head scatter factor was measured using a mini-phantom and the effect of leaf guide position on output was assessed using a water phantom. The tongue and groove effect was assessed using multiple exposures on radiochromic film. Leaf reproducibility was assessed from portal images of multiple abutting fields. Results: The maximum transmission through the multileaf collimator is 0.44% at 6 MV and 0.52% at 10 MV. This reduced to 0.22% and 0.27%, respectively, when the beam passes through the dynamic leaf guide in addition to the MLC. The maximum transmission through the thick part of the diaphragm is 0.32% and 0.36% at 6 and 10 MV. The combination of leaf and diaphragm transmission ranges from 0.08% to 0.010% at 6 MV and 0.10% to 0.14% depending on whether the shielding is through the thick or thin part of the diaphragm. The off-axis intertip transmission for a zero leaf gap is 2.2% at 6 and 10 MV. The leaf tip penumbra for a 100 × 100 mm field ranges from 5.4 to 4.3 mm at 6 and 10 MV across the full range of leaf motion when measured in the AB direction, which reduces to 4.0–3.4 mm at 6 MV and 4.5–3.8 mm at 10 MV when measured in the GT direction. For a 50 × 50 mm field, the diaphragm penumbra ranges from 4.3 to 3.7 mm at 6 MV and 4.5 to 4.1 mm at 10 MV in the AB direction and 3.7 to 3.2 mm at 6 MV and 4.2 to 3.7 mm when measured in the GT direction. The

  8. Feasibility of a simple method of hybrid collimation for megavoltage grid therapy

    SciTech Connect

    Almendral, Pedro; Mancha, Pedro J.; Roberto, Daniel

    2013-05-15

    Purpose: Megavoltage grid therapy is currently delivered with step-and-shoot multisegment techniques or using a high attenuation block with divergent holes. However, the commercial availability of grid blocks is limited, their construction is difficult, and step-and-shoot techniques require longer treatment times and are not practical with some multileaf collimators. This work studies the feasibility of a hybrid collimation system for grid therapy that does not require multiple segments and can be easily implemented with widely available technical means. Methods: The authors have developed a system to generate a grid of beamlets by the simultaneous use of two perpendicular sets of equally spaced leaves that project stripe patterns in orthogonal directions. One of them is generated with the multileaf collimator integrated in the accelerator and the other with an in-house made collimator constructed with a low melting point alloy commonly available at radiation oncology departments. The characteristics of the grid fields for 6 and 18 MV have been studied with a shielded diode, an unshielded diode, and radiochromic film. Results: The grid obtained with the hybrid collimation is similar to some of the grids used clinically with respect to the beamlet size (about 1 cm) and the percentage of open beam (1/4 of the total field). The grid fields are less penetrating than the open fields of the same energy. Depending on the depth and the direction of the profiles (diagonal or along the principal axes), the measured valley-to-peak dose ratios range from 5% to 16% for 6 MV and from 9% to 20% for 18 MV. All the detectors yield similar results in the measurement of profiles and percent depth dose, but the shielded diode seems to overestimate the output factors. Conclusions: The combination of two stripe pattern collimators in orthogonal directions is a feasible method to obtain two-dimensional arrays of beamlets and has potential usefulness as an efficient way to deliver grid

  9. Optimization of the rounded leaf offset table in modeling the multileaf collimator leaf edge in a commercial treatment planning system.

    PubMed

    Rice, John R

    2014-11-08

    An editable rounded leaf offset (RLO) table is provided in the Pinnacle3 treatment planning software. Default tables are provided for major linear accelerator manu- facturers, but it is not clear how the default table values should be adjusted by the user to optimize agreement between the calculated leaf tip value and the actual measured value. Since we wish for the calculated MLC-defined field edge to closely match the actual delivered field edge, optimal RLO table values are crucial. This is especially true for IMRT fields containing a large number of segments, since any errors would add together. A method based on the calculated MLC-defined field edge was developed for optimizing and modifying the default RLO table values. Modified RLO tables were developed and evaluated for both dosimetric and light field-based MLC leaf calibrations. It was shown, using a Picket Fence type test, that the optimized RLO table better modeled the calculated leaf tip than the Pinnacle3 default table. This was demonstrated for both an Elekta Synergy 80-leaf and a Varian 120-leaf MLC. 

  10. Wakefields in SLAC linac collimators

    NASA Astrophysics Data System (ADS)

    Novokhatski, A.; Decker, F.-J.; Smith, H.; Sullivan, M.

    2014-12-01

    When a beam travels near collimator jaws, it gets an energy loss and a transverse kick due to the backreaction of the beam field diffracted from the jaws. The effect becomes very important for an intense short bunch when a tight collimation of the background beam halo is required. In the Linac Coherent Light Source at SLAC a collimation system is used to protect the undulators from radiation due to particles in the beam halo. The halo is most likely formed from gun dark current or dark current in some of the accelerating sections. However, collimators are also responsible for the generation of wake fields. The wake field effect from the collimators not only brings an additional energy jitter and change in the trajectory of the beam, but it also rotates the beam on the phase plane, which consequently leads to a degradation of the performance of the Free Electron Laser at the Linac Coherent Light Source. In this paper, we describe a model of the wake field radiation in the SLAC linac collimators. We use the results of a numerical simulation to illustrate the model. Based on the model, we derive simple formulas for the bunch energy loss and the average kick. We also present results from experimental measurements that confirm our model.

  11. Design of multipinhole collimators for small animal SPECT

    SciTech Connect

    Mark Smith; Steve Meikle; Stanislaw Majewski; Andrew Weisenberger

    2003-10-01

    High resolution single photon emission computed tomography (SPECT) of small animals with pinhole collimators is hampered by poor sensitivity. For the target application of imaging the brain of an unanesthetized, unrestrained mouse, the design of multipinhole collimators for small, compact gamma cameras was studied using numerical simulations. The collimators were designed to image the mouse at different axial positions inside a tube aligned with the axis of rotation (AOR) of the SPECT system. The proposed system consisted of a 10 cm/spl times/20 cm pixelated detector with 2 cm from the AOR to the collimator mask and a focal length of 7 cm for a magnification of 3.5 on the AOR. A digital mouse phantom was created with a brain:background activity concentration ratio of 8:1. The brain region contained small disks and hot and cool cylinders for contrast and signal-to-noise ratio (SNR) quantitation. SPECT projection data were simulated at 120 angles for four pinhole masks having 1, 5 and two arrangements of 13 pinholes. Projections were scaled to model a 30 min study with doses of 0.3, 1 and 3 mCi in the mouse. Images were reconstructed with an iterative ordered subset expectation maximization (OSEM) algorithm. The peak SNR improved with multipinhole collimation compared with single pinhole imaging when there was no increased activity in the liver or bladder. With such activity, peak SNR was about the same for all collimators. For all cases peak SNR occurred at higher contrast values with multipinhole collimation. The extended axial range of multipinhole arrays enabled the mouse brain to be imaged with decreased axial blurring when the mouse was translated axially. These results suggest that mouse brains can be successfully imaged at realistic activity levels with multipinhole arrays and projection data multiplexing.

  12. Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission

    PubMed Central

    Didi, Samir; Moussa, Abdelilah; Yahya, Tayalati; Mustafa, Zerfaoui

    2015-01-01

    The present work validates the Geant4 Application for Tomographic Emission Monte Carlo software for the simulation of a 6 MV photon beam given by Elekta Synergy Platform medical linear accelerator treatment head. The simulation includes the major components of the linear accelerator (LINAC) with multi-leaf collimator and a homogeneous water phantom. Calculations were performed for the photon beam with several treatment field sizes ranging from 5 cm × 5 cm to 30 cm × 30 cm at 100 cm distance from the source. The simulation was successfully validated by comparison with experimental distributions. Good agreement between simulations and measurements was observed, with dose differences of about 0.02% and 2.5% for depth doses and lateral dose profiles, respectively. This agreement was also emphasized by the Kolmogorov–Smirnov goodness-of-fit test and by the gamma-index comparisons where more than 99% of the points for all simulations fulfill the quality assurance criteria of 2 mm/2%. PMID:26500399

  13. SU-E-T-11: A Dosimetric Comparison of Robotic Prostatic Radiosugery Using Multi- Leaf Collimation Vs Circular Collimators

    SciTech Connect

    Feng, J; Yang, J; Lamond, J; Lavere, N; Laciano, R; Ding, W; Arrigo, S; Brady, L

    2014-06-01

    Purpose: The study compared the dosimetry plans of Stereotatic Body Radiotherapy (SBRT) prostate cancer patients using the M6 Cyberknife with Multi-leaf Collimation (MLC) compared with the plans using G4 Cyberknife with circular collimators. Methods: Eight previously treated prostate cancer patients' SBRT plans using circular collimators, designed with Multiplan v3.5.3, were used as a benchmark. The CT, contours and the optimization scripts were imported into Multiplan v5.0 system and replanned with MLC. The same planning objectives were used: more than 95% of PTV received 36.25Gy, 90% of prostate received 40Gy and maximum dose <45Gy, in five fractions. For organs at risk, less than 1cc of rectum received 36Gy and less than 10cc of bladder received 37Gy. Plans were evaluated on parameters derived from dose volume. The beam number, MU and delivery time were recorded to compare the treatment efficiency. Results: The mean CTV volume was 41.3cc (27.5∼57.6cc) and mean PTV volume was 76.77cc (59.1∼99.7cc). The mean PTV coverage was comparable between MLC (98.87%) and cone (98.74%). MLC plans had a slightly more favorable homogeneity index (1.22) and conformity index (1.17), than the cone (1.24 and 1.15). The mean rectum volume of 36 Gy (0.52cc) of MLC plans was slightly larger than cone (0.38cc) and the mean bladder volume of 37 Gy was smaller in MLC (1.82cc) than in cone plans (3.09cc). The mean number of nodes and beams were 65.9 and 80.5 in MLC vs 65.9 and 203.6 in cone. The mean MUs were significantly less for MLC plans (24,228MUs) than cone (32,347MUs). The total delivery time (which included 5 minutes for setup) was less, 29.6min (26∼32min) for MLC vs 45min (35∼55min) for cone. Conclusion: While the differences in the dosimetry between the MLC and circular collimator plans were rather minor, the MLC plans were much more efficient and required significantly less treatment time.

  14. Optimizing Pinhole and Parallel Hole Collimation for Scintimammography with Compact Pixellated Detectors

    SciTech Connect

    Mark F. Smith; Douglas A. Kieper; Stan Majewski; Andrew G. Weisenberger; Benjamin Welch

    2002-11-01

    The relative advantages of pinhole and parallel hole collimators for scintimammography with compact, pixellated gamma detectors were investigated using analytic models of resolution and sensitivity. Collimator design was studied as follows. A desired object resolution was specified for a pixellated detector with a given crystal size and intrinsic spatial resolution and for a given object-to-collimator distance. Using analytic formulas, pinhole and parallel hole collimator parameters were calculated that satisfy this object resolution with optimal geometric sensitivity. Analyses were performed for 15 cm x 20 cm field of view detectors with crystal elements 1.0, 2.0 and 3.0 mm on a side and 140 keV incident photons. The sensitivity for a given object resolution was greater for pinhole collimation at smaller distances, as expected. The object distance at which the pinhole and parallel hole sensitivity curves cross each other is important. The crossover distances increased with larger crystal size for a constant object resolution and increased as the desired object resolution decreases for a constant crystal size. For example, for 4 mm object resolution these distances were 5.5 cm, 6.5 cm and 8 cm for the 1 mm, 2 mm and 3 mm crystal detectors, respectively. The results suggest a strategy of parallel hole collimation for whole breast imaging and pinhole collimation for imaging focal uptake. This could be accomplished with a dual detector system with a different collimator type on each head or a single head system equipped with two collimators and a rapid switching mechanism. Multipinhole collimators have the potential to increase sensitivity yet maintain high image resolution. An experimental SPECT phantom study with a four-pinhole collimator was acquired with a pixellated detector. The iterative maximum-likelihood expectation- maximization (MLEM) reconstruction of a hot sphere in a warm cylinder showed the potential of multipinhole collimation to improve sensitivity

  15. Surface Dose Investigation of the Flattening Filter-Free Photon Beams

    SciTech Connect

    Wang Yuenan; Khan, Mohammad K.; Ting, Joseph Y.; Easterling, Stephen B.

    2012-06-01

    Purpose: Flattening filter-free (FFF) x-rays can provide more efficient use of photons and a significant increase of dose rate compared with conventional flattened x-rays, features that are especially beneficial for stereotactic radiosurgery (SRS) and stereotactic body radiotherapy (SBRT). The available data on the entrance doses of the FFF photon beams remain limited. The purpose of this study was to investigate the entrance dose of FFF photons in the buildup region and to compare it with that of conventional flattened photons. Methods and Materials: A Varian TrueBeam linear accelerator has been in full clinical operation with 6-MV and 10-MV FFF and flattened x-ray photons. Entrance dose at the surface was measured using a parallel plate ionization chamber in a solid water phantom with buildup depth = 0{approx}15 mm for 6X and 0{approx}25 mm for 10X. Different field size (FS) patterns were created in the Eclipse Treatment Planning System by multileaf collimator (MLC) rather than jaws (FS = 2 Multiplication-Sign 2, 3 Multiplication-Sign 3, 4 Multiplication-Sign 4, 6 Multiplication-Sign 6, and 10 Multiplication-Sign 10 cm{sup 2} by MLC and jaw size = 2.2 Multiplication-Sign 2.2, 3.2 Multiplication-Sign 3.2, 4.2 Multiplication-Sign 4.2, 6 Multiplication-Sign 6, and 10 Multiplication-Sign 10 cm{sup 2}). The smallest FS was about four times larger than the ion chamber dimension. All buildup dose measurements were normalized to FS = 10 Multiplication-Sign 10 cm{sup 2} at the depth of dose maximum (dmax). Results: Good repeatability was demonstrated and surface dose increased linearly with FS for both flattened and FFF photons. The entrance dose of the FFF photons was modestly larger than that of the corresponding flattened photons for both 6X and 10X for different FS ranging from 2 Multiplication-Sign 2 cm{sup 2} to 10 Multiplication-Sign 10 cm{sup 2}. Conclusions: The FFF photons have a higher entrance dose than that of the corresponding flattened photons for FS

  16. Incorporating dynamic collimator motion in Monte Carlo simulations: an application in modelling a dynamic wedge

    NASA Astrophysics Data System (ADS)

    Verhaegen, Frank; Liu, H. Helen

    2001-02-01

    In radiation therapy, new treatment modalities employing dynamic collimation and intensity modulation increase the complexity of dose calculation because a new dimension, time, has to be incorporated into the traditional three-dimensional problem. In this work, we investigated two classes of sampling technique to incorporate dynamic collimator motion in Monte Carlo simulation. The methods were initially evaluated for modelling enhanced dynamic wedges (EDWs) from Varian accelerators (Varian Medical Systems, Palo Alto, USA). In the position-probability-sampling or PPS method, a cumulative probability distribution function (CPDF) was computed for the collimator position, which could then be sampled during simulations. In the static-component-simulation or SCS method, a dynamic field is approximated by multiple static fields in a step-shoot fashion. The weights of the particles or the number of particles simulated for each component field are computed from the probability distribution function (PDF) of the collimator position. The CPDF and PDF were computed from the segmented treatment tables (STTs) for the EDWs. An output correction factor had to be applied in this calculation to account for the backscattered radiation affecting monitor chamber readings. Comparison of the phase-space data from the PPS method (with the step-shoot motion) with those from the SCS method showed excellent agreement. The accuracy of the PPS method was further verified from the agreement between the measured and calculated dose distributions. Compared to the SCS method, the PPS method is more automated and efficient from an operational point of view. The principle of the PPS method can be extended to simulate other dynamic motions, and in particular, intensity-modulated beams using multileaf collimators.

  17. Beam collimation at hadron colliders

    SciTech Connect

    Nikolai V. Mokhov

    2003-08-12

    Operational and accidental beam losses in hadron colliders can have a serious impact on machine and detector performance, resulting in effects ranging from minor to catastrophic. Principles and realization are described for a reliable beam collimation system required to sustain favorable background conditions in the collider detectors, provide quench stability of superconducting magnets, minimize irradiation of accelerator equipment, maintain operational reliability over the life of the machine, and reduce the impact of radiation on personnel and the environment. Based on detailed Monte-Carlo simulations, such a system has been designed and incorporated in the Tevatron collider. Its performance, comparison to measurements and possible ways to further improve the collimation efficiency are described in detail. Specifics of the collimation systems designed for the SSC, LHC, VLHC, and HERA colliders are discussed.

  18. Effect of MLC leaf position, collimator rotation angle, and gantry rotation angle errors on intensity-modulated radiotherapy plans for nasopharyngeal carcinoma

    SciTech Connect

    Bai, Sen; Li, Guangjun; Wang, Maojie; Jiang, Qinfeng; Zhang, Yingjie; Wei, Yuquan

    2013-07-01

    The purpose of this study was to investigate the effect of multileaf collimator (MLC) leaf position, collimator rotation angle, and accelerator gantry rotation angle errors on intensity-modulated radiotherapy plans for nasopharyngeal carcinoma. To compare dosimetric differences between the simulating plans and the clinical plans with evaluation parameters, 6 patients with nasopharyngeal carcinoma were selected for simulation of systematic and random MLC leaf position errors, collimator rotation angle errors, and accelerator gantry rotation angle errors. There was a high sensitivity to dose distribution for systematic MLC leaf position errors in response to field size. When the systematic MLC position errors were 0.5, 1, and 2 mm, respectively, the maximum values of the mean dose deviation, observed in parotid glands, were 4.63%, 8.69%, and 18.32%, respectively. The dosimetric effect was comparatively small for systematic MLC shift errors. For random MLC errors up to 2 mm and collimator and gantry rotation angle errors up to 0.5°, the dosimetric effect was negligible. We suggest that quality control be regularly conducted for MLC leaves, so as to ensure that systematic MLC leaf position errors are within 0.5 mm. Because the dosimetric effect of 0.5° collimator and gantry rotation angle errors is negligible, it can be concluded that setting a proper threshold for allowed errors of collimator and gantry rotation angle may increase treatment efficacy and reduce treatment time.

  19. Neutron collimator for neutron radiography applications at tangential port of the TRIGA RC-1 reactor

    NASA Astrophysics Data System (ADS)

    Rosa, R.; Andreoli, F.; Mattoni, M.; Palomba, M.

    2009-06-01

    At the ENEA TRIGA research reactor (Casaccia Research Center, Rome) a new neutron collimator has been designed and installed at the neutron tangential channel. This collimator, that is part of a neutron/X-ray facility for NDT analysis, was experimentally characterized and optimized in terms of thermal neutron fluence rate, spatial/energetic distribution, photon air KERMA and effective beam diameter. This paper shows the methodologies and the results of the experimental analysis that were carried out.

  20. [Dual copy-supplement collimator].

    PubMed

    Wang, Z K; Lian, X; Wang, Y

    2001-01-01

    Dual Copy-Supplement Collimator consists of two lead conic sections centered on one common radioactive source and having a lot of small square holes on them. When scanning electron beams strike X-Ray target and create some X-Ray fields while passing through the primary section, they penetrate the secondary section and generate even smaller field units(FU). Using these FUs would compose various treatment fields with different energy (dosage). It is able not only to replace many existing collimators such as MLC, but also to make the conformal radiotherapy easier.

  1. Optimizing Collimator Margins for Isotoxically Dose-Escalated Conformal Radiation Therapy of Non-Small Cell Lung Cancer

    SciTech Connect

    Warren, Samantha; Panettieri, Vanessa; Panakis, Niki; Bates, Nicholas; Lester, Jason F.; Jain, Pooja; Landau, David B.; Nahum, Alan E.; Mayles, W. Philip M.; Fenwick, John D.

    2014-04-01

    Purpose: Isotoxic dose escalation schedules such as IDEAL-CRT [isotoxic dose escalation and acceleration in lung cancer chemoradiation therapy] (ISRCTN12155469) individualize doses prescribed to lung tumors, generating a fixed modeled risk of radiation pneumonitis. Because the beam penumbra is broadened in lung, the choice of collimator margin is an important element of the optimization of isotoxic conformal radiation therapy for lung cancer. Methods and Materials: Twelve patients with stage I-III non-small cell lung cancer (NSCLC) were replanned retrospectively using a range of collimator margins. For each plan, the prescribed dose was calculated according to the IDEAL-CRT isotoxic prescription method, and the absolute dose (D{sub 99}) delivered to 99% of the planning target volume (PTV) was determined. Results: Reducing the multileaf collimator margin from the widely used 7 mm to a value of 2 mm produced gains of 2.1 to 15.6 Gy in absolute PTV D{sub 99}, with a mean gain ± 1 standard error of the mean of 6.2 ± 1.1 Gy (2-sided P<.001). Conclusions: For NSCLC patients treated with conformal radiation therapy and an isotoxic dose prescription, absolute doses in the PTV may be increased by using smaller collimator margins, reductions in relative coverage being offset by increases in prescribed dose.

  2. Characterization of parallel-hole collimator using Monte Carlo Simulation

    PubMed Central

    Pandey, Anil Kumar; Sharma, Sanjay Kumar; Karunanithi, Sellam; Kumar, Praveen; Bal, Chandrasekhar; Kumar, Rakesh

    2015-01-01

    Objective: Accuracy of in vivo activity quantification improves after the correction of penetrated and scattered photons. However, accurate assessment is not possible with physical experiment. We have used Monte Carlo Simulation to accurately assess the contribution of penetrated and scattered photons in the photopeak window. Materials and Methods: Simulations were performed with Simulation of Imaging Nuclear Detectors Monte Carlo Code. The simulations were set up in such a way that it provides geometric, penetration, and scatter components after each simulation and writes binary images to a data file. These components were analyzed graphically using Microsoft Excel (Microsoft Corporation, USA). Each binary image was imported in software (ImageJ) and logarithmic transformation was applied for visual assessment of image quality, plotting profile across the center of the images and calculating full width at half maximum (FWHM) in horizontal and vertical directions. Results: The geometric, penetration, and scatter at 140 keV for low-energy general-purpose were 93.20%, 4.13%, 2.67% respectively. Similarly, geometric, penetration, and scatter at 140 keV for low-energy high-resolution (LEHR), medium-energy general-purpose (MEGP), and high-energy general-purpose (HEGP) collimator were (94.06%, 3.39%, 2.55%), (96.42%, 1.52%, 2.06%), and (96.70%, 1.45%, 1.85%), respectively. For MEGP collimator at 245 keV photon and for HEGP collimator at 364 keV were 89.10%, 7.08%, 3.82% and 67.78%, 18.63%, 13.59%, respectively. Conclusion: Low-energy general-purpose and LEHR collimator is best to image 140 keV photon. HEGP can be used for 245 keV and 364 keV; however, correction for penetration and scatter must be applied if one is interested to quantify the in vivo activity of energy 364 keV. Due to heavy penetration and scattering, 511 keV photons should not be imaged with HEGP collimator. PMID:25829730

  3. Proton Collimators for Fusion Reactors

    NASA Technical Reports Server (NTRS)

    Miley, George H.; Momota, Hiromu

    2003-01-01

    Proton collimators have been proposed for incorporation into inertial-electrostatic-confinement (IEC) fusion reactors. Such reactors have been envisioned as thrusters and sources of electric power for spacecraft and as sources of energetic protons in commercial ion-beam applications.

  4. Collimator with attachment mechanism and system

    DOEpatents

    Kross, Brian J.; McKisson, John; Stolin, Aleksandr; Weisenberger, Andrew G.; Zorn, Carl

    2012-07-10

    A self-aligning collimator for a radiation imaging device that is secured and aligned through the use of a plurality of small magnets. The collimator allows for the rapid exchange, removal, or addition of collimators for the radiation imaging device without the need for tools. The accompanying method discloses the use of magnets and accompanying magnetic fields to align and secure collimators in a radiation imaging assembly.

  5. Carbon nanotube collimator fabrication and application

    DOEpatents

    Chow, Lee; Chai, Guangyu; Schenkel, Thomas

    2010-07-06

    Apparatus, methods, systems and devices for fabricating individual CNT collimators. Micron size fiber coated CNT samples are synthesized with chemical vapor deposition method and then the individual CNT collimators are fabricated with focused ion beam technique. Unfocused electron beams are successfully propagated through the CNT collimators. The CNT nano-collimators are used for applications including single ion implantation and in high-energy physics, and allow rapid, reliable testing of the transmission of CNT arrays for transport of molecules.

  6. High resolution alpha particle spectrometry through collimation

    NASA Astrophysics Data System (ADS)

    Park, Seunghoon; Kwak, Sung-Woo; Kang, Han-Byeol

    2015-06-01

    Alpha particle spectrometry with collimation is a useful method for identifying nuclear materials among various nuclides. A mesh type collimator reduces the low energy tail and broadened energy distribution by cutting off particles with a low incidence angle. The relation between the resolution and the counting efficiency can be investigated by changing a ratio of the mesh hole diameter and the collimator thickness. Through collimation, a target particle can be distinguished by a PIPS® detector under a mixture of various nuclides.

  7. Monitor unit calculations for external photon and electron beams: Report of the AAPM Therapy Physics Committee Task Group No. 71.

    PubMed

    Gibbons, John P; Antolak, John A; Followill, David S; Huq, M Saiful; Klein, Eric E; Lam, Kwok L; Palta, Jatinder R; Roback, Donald M; Reid, Mark; Khan, Faiz M

    2014-03-01

    A protocol is presented for the calculation of monitor units (MU) for photon and electron beams, delivered with and without beam modifiers, for constant source-surface distance (SSD) and source-axis distance (SAD) setups. This protocol was written by Task Group 71 of the Therapy Physics Committee of the American Association of Physicists in Medicine (AAPM) and has been formally approved by the AAPM for clinical use. The protocol defines the nomenclature for the dosimetric quantities used in these calculations, along with instructions for their determination and measurement. Calculations are made using the dose per MU under normalization conditions, D'0, that is determined for each user's photon and electron beams. For electron beams, the depth of normalization is taken to be the depth of maximum dose along the central axis for the same field incident on a water phantom at the same SSD, where D'0 = 1 cGy/MU. For photon beams, this task group recommends that a normalization depth of 10 cm be selected, where an energy-dependent D'0 ≤ 1 cGy/MU is required. This recommendation differs from the more common approach of a normalization depth of dm, with D'0 = 1 cGy/MU, although both systems are acceptable within the current protocol. For photon beams, the formalism includes the use of blocked fields, physical or dynamic wedges, and (static) multileaf collimation. No formalism is provided for intensity modulated radiation therapy calculations, although some general considerations and a review of current calculation techniques are included. For electron beams, the formalism provides for calculations at the standard and extended SSDs using either an effective SSD or an air-gap correction factor. Example tables and problems are included to illustrate the basic concepts within the presented formalism.

  8. Monitor unit calculations for external photon and electron beams: Report of the AAPM Therapy Physics Committee Task Group No. 71

    SciTech Connect

    Gibbons, John P.; Antolak, John A.; Followill, David S.; Huq, M. Saiful; Klein, Eric E.; Lam, Kwok L.; Palta, Jatinder R.; Roback, Donald M.; Reid, Mark; Khan, Faiz M.

    2014-03-15

    A protocol is presented for the calculation of monitor units (MU) for photon and electron beams, delivered with and without beam modifiers, for constant source-surface distance (SSD) and source-axis distance (SAD) setups. This protocol was written by Task Group 71 of the Therapy Physics Committee of the American Association of Physicists in Medicine (AAPM) and has been formally approved by the AAPM for clinical use. The protocol defines the nomenclature for the dosimetric quantities used in these calculations, along with instructions for their determination and measurement. Calculations are made using the dose per MU under normalization conditions, D{sub 0}{sup ′}, that is determined for each user's photon and electron beams. For electron beams, the depth of normalization is taken to be the depth of maximum dose along the central axis for the same field incident on a water phantom at the same SSD, where D{sub 0}{sup ′} = 1 cGy/MU. For photon beams, this task group recommends that a normalization depth of 10 cm be selected, where an energy-dependent D{sub 0}{sup ′} ≤ 1 cGy/MU is required. This recommendation differs from the more common approach of a normalization depth of d{sub m}, with D{sub 0}{sup ′} = 1 cGy/MU, although both systems are acceptable within the current protocol. For photon beams, the formalism includes the use of blocked fields, physical or dynamic wedges, and (static) multileaf collimation. No formalism is provided for intensity modulated radiation therapy calculations, although some general considerations and a review of current calculation techniques are included. For electron beams, the formalism provides for calculations at the standard and extended SSDs using either an effective SSD or an air-gap correction factor. Example tables and problems are included to illustrate the basic concepts within the presented formalism.

  9. Development of a new, fast, user friendly, ray tracing program "CSIM" for the simulation of parallelhole collimators.

    PubMed

    Erturk, Sakine Sebnem; Del Guerra, Alberto

    2013-06-01

    We have developed a fast, user friendly, ray-tracing program, "CSIM" for low-energy gamma rays (up to ∼200keV) to simulate the performance characteristics of parallelhole collimators. We have used a ray-tracing approach to find the sensitivity and resolution of the parallelhole collimator by including the penetration of photons through the collimator due to the finite attenuation of the collimator material. "CSIM" can calculate the sensitivity of the collimator, the geometric and penetrating photon ratios, and the 1D and 2D point source response functions (PSF) with the statistical uncertainty for different hole shapes (e.g. square, hexagonal, and cylindrical). We have used "CSIM" to simulate the collimator of the YAP-(S)PETII small animal scanner. We present the analysis of the YAP-(S)PETII scanner round-hole parallel collimator designed for nuclear medicine imaging at 140keV. For this aim, different designs have been considered for a variety of source-collimator distances (b=5, 10, 15, 20cm). Resolution and sensitivity characteristics have been plotted as a function of the collimator thickness and the diameter of the hole. For each value of the source-collimator distance, and for each collimator thickness investigated, the trade-off between sensitivity and spatial resolution has been given as a series of characteristic curves. Then, we compare our simulated resolution and sensitivity results to the analytically calculated ones and found that the analytically calculated results for the YAP-(S)PETII scanner collimator are not far away the results predicted by CSIM and also with the experimentally measured resolution values.

  10. Evaluation of uncertainty predictions and dose output for model-based dose calculations for megavoltage photon beams

    SciTech Connect

    Olofsson, Joergen; Nyholm, Tufve; Georg, Dietmar; Ahnesjoe, Anders; Karlsson, Mikael

    2006-07-15

    In many radiotherapy clinics an independent verification of the number of monitor units (MU) used to deliver the prescribed dose to the target volume is performed prior to the treatment start. Traditionally this has been done by using methods mainly based on empirical factors which, at least to some extent, try to separate the influence from input parameters such as field size, depth, distance, etc. The growing complexity of modern treatment techniques does however make this approach increasingly difficult, both in terms of practical application and in terms of the reliability of the results. In the present work the performance of a model-based approach, describing the influence from different input parameters through actual modeling of the physical effects, has been investigated in detail. The investigated model is based on two components related to megavoltage photon beams; one describing the exiting energy fluence per delivered MU, and a second component describing the dose deposition through a pencil kernel algorithm solely based on a measured beam quality index. Together with the output calculations, the basis of a method aiming to predict the inherent calculation uncertainties in individual treatment setups has been developed. This has all emerged from the intention of creating a clinical dose/MU verification tool that requires an absolute minimum of commissioned input data. This evaluation was focused on irregular field shapes and performed through comparison with output factors measured at 5, 10, and 20 cm depth in ten multileaf collimated fields on four different linear accelerators with varying multileaf collimator designs. The measurements were performed both in air and in water and the results of the two components of the model were evaluated separately and combined. When compared with the corresponding measurements the resulting deviations in the calculated output factors were in most cases smaller than 1% and in all cases smaller than 1.7%. The

  11. Evaluation of uncertainty predictions and dose output for model-based dose calculations for megavoltage photon beams.

    PubMed

    Olofsson, Jörgen; Nyholm, Tufve; Georg, Dietmar; Ahnesjö, Anders; Karlsson, Mikael

    2006-07-01

    In many radiotherapy clinics an independent verification of the number of monitor units (MU) used to deliver the prescribed dose to the target volume is performed prior to the treatment start. Traditionally this has been done by using methods mainly based on empirical factors which, at least to some extent, try to separate the influence from input parameters such as field size, depth, distance, etc. The growing complexity of modern treatment techniques does however make this approach increasingly difficult, both in terms of practical application and in terms of the reliability of the results. In the present work the performance of a model-based approach, describing the influence from different input parameters through actual modeling of the physical effects, has been investigated in detail. The investigated model is based on two components related to megavoltage photon beams; one describing the exiting energy fluence per delivered MU, and a second component describing the dose deposition through a pencil kernel algorithm solely based on a measured beam quality index. Together with the output calculations, the basis of a method aiming to predict the inherent calculation uncertainties in individual treatment setups has been developed. This has all emerged from the intention of creating a clinical dose/MU verification tool that requires an absolute minimum of commissioned input data. This evaluation was focused on irregular field shapes and performed through comparison with output factors measured at 5, 10, and 20 cm depth in ten multileaf collimated fields on four different linear accelerators with varying multileaf collimator designs. The measurements were performed both in air and in water and the results of the two components of the model were evaluated separately and combined. When compared with the corresponding measurements the resulting deviations in the calculated output factors were in most cases smaller than 1% and in all cases smaller than 1.7%. The

  12. Optimization of the CLIC Baseline Collimation System

    SciTech Connect

    Resta-Lopez, Javier; Angal-Kalinin, Deepa; Fernandez-Hernando, Juan; Jackson, Frank; Dalena, Barbara; Schulte, Daniel; Tomas, Rogelio; Seryi, Andrei; /SLAC

    2012-07-06

    Important efforts have recently been dedicated to the improvement of the design of the baseline collimation system of the Compact Linear Collider (CLIC). Different aspects of the design have been optimized: the transverse collimation depths have been recalculated in order to reduce the collimator wakefield effects while maintaining a good efficiency in cleaning the undesired beam halo; the geometric design of the spoilers have also been reviewed to minimize wakefields; in addition, the optics design have been polished to improve the collimation efficiency. This paper describes the current status of the CLIC collimation system after this optimization.

  13. Optimizing Pinhole and Parallel Hole Collimation for Scintimammography With Compact Pixellated Detectors

    SciTech Connect

    Mark F. Smith; Stan Majewski; Andrew G. Weisenberger

    2002-11-01

    The relative resolution and sensitivity advantages of pinhole and parallel hole collimators for planar scintimammography with compact, pixellated gamma detectors were investigated using analytic models. Collimator design was studied as follows. A desired object resolution was specified for a pixellated detector with a given crystal size and intrinsic spatial resolution and for a given object-to- collimator distance. Using analytic formulas, pinhole and parallel hole collimator parameters were calculated that satisfy this object resolution with optimal geometric sensitivity. Analyses were performed for 15 cm x 20 cm field of view detectors with crystal elements 1.0, 2.0 and 3.0 mm on a side and 140 keV incident photons. The sensitivity for a given object resolution was greater for pinhole collimation at smaller distances, as expected. The object distance at which the pinhole and parallel hole sensitivity curves cross each other is important. The crossover distances increased with larger crystal size for a constant object resolution and increased as the desired object resolution decreases for a constant crystal size. For example, for 4 mm object resolution and a pinhole collimator with focal length 13 cm, these distances were 5.5 cm, 6.5 cm and 8 cm for the 1 mm, 2 mm and 3 mm crystal detectors, respectively. The results suggest a strategy of parallel hole collimation for whole breast imaging and pinhole collimation for imaging focal uptake. This could be accomplished with a dual detector system with a different collimator type on each head or a single head system equipped with two collimators and a rapid switching mechanism.

  14. Fabricating High-Resolution X-Ray Collimators

    NASA Technical Reports Server (NTRS)

    Appleby, Michael; Atkinson, James E.; Fraser, Iain; Klinger, Jill

    2008-01-01

    A process and method for fabricating multi-grid, high-resolution rotating modulation collimators for arcsecond and sub-arcsecond x-ray and gamma-ray imaging involves photochemical machining and precision stack lamination. The special fixturing and etching techniques that have been developed are used for the fabrication of multiple high-resolution grids on a single array substrate. This technology has application in solar and astrophysics and in a number of medical imaging applications including mammography, computed tomography (CT), single photon emission computed tomography (SPECT), and gamma cameras used in nuclear medicine. This collimator improvement can also be used in non-destructive testing, hydrodynamic weapons testing, and microbeam radiation therapy.

  15. Multiple pinhole collimator based X-ray luminescence computed tomography.

    PubMed

    Zhang, Wei; Zhu, Dianwen; Lun, Michael; Li, Changqing

    2016-07-01

    X-ray luminescence computed tomography (XLCT) is an emerging hybrid imaging modality, which is able to improve the spatial resolution of optical imaging to hundreds of micrometers for deep targets by using superfine X-ray pencil beams. However, due to the low X-ray photon utilization efficiency in a single pinhole collimator based XLCT, it takes a long time to acquire measurement data. Herein, we propose a multiple pinhole collimator based XLCT, in which multiple X-ray beams are generated to scan a sample at multiple positions simultaneously. Compared with the single pinhole based XLCT, the multiple X-ray beam scanning method requires much less measurement time. Numerical simulations and phantom experiments have been performed to demonstrate the feasibility of the multiple X-ray beam scanning method. In one numerical simulation, we used four X-ray beams to scan a cylindrical object with 6 deeply embedded targets. With measurements from 6 angular projections, all 6 targets have been reconstructed successfully. In the phantom experiment, we generated two X-ray pencil beams with a collimator manufactured in-house. Two capillary targets with 0.6 mm edge-to-edge distance embedded in a cylindrical phantom have been reconstructed successfully. With the two beam scanning, we reduced the data acquisition time by 50%. From the reconstructed XLCT images, we found that the Dice similarity of targets is 85.11% and the distance error between two targets is less than 3%. We have measured the radiation dose during XLCT scan and found that the radiation dose, 1.475 mSv, is in the range of a typical CT scan. We have measured the changes of the collimated X-ray beam size and intensity at different distances from the collimator. We have also studied the effects of beam size and intensity in the reconstruction of XLCT.

  16. Multiple pinhole collimator based X-ray luminescence computed tomography.

    PubMed

    Zhang, Wei; Zhu, Dianwen; Lun, Michael; Li, Changqing

    2016-07-01

    X-ray luminescence computed tomography (XLCT) is an emerging hybrid imaging modality, which is able to improve the spatial resolution of optical imaging to hundreds of micrometers for deep targets by using superfine X-ray pencil beams. However, due to the low X-ray photon utilization efficiency in a single pinhole collimator based XLCT, it takes a long time to acquire measurement data. Herein, we propose a multiple pinhole collimator based XLCT, in which multiple X-ray beams are generated to scan a sample at multiple positions simultaneously. Compared with the single pinhole based XLCT, the multiple X-ray beam scanning method requires much less measurement time. Numerical simulations and phantom experiments have been performed to demonstrate the feasibility of the multiple X-ray beam scanning method. In one numerical simulation, we used four X-ray beams to scan a cylindrical object with 6 deeply embedded targets. With measurements from 6 angular projections, all 6 targets have been reconstructed successfully. In the phantom experiment, we generated two X-ray pencil beams with a collimator manufactured in-house. Two capillary targets with 0.6 mm edge-to-edge distance embedded in a cylindrical phantom have been reconstructed successfully. With the two beam scanning, we reduced the data acquisition time by 50%. From the reconstructed XLCT images, we found that the Dice similarity of targets is 85.11% and the distance error between two targets is less than 3%. We have measured the radiation dose during XLCT scan and found that the radiation dose, 1.475 mSv, is in the range of a typical CT scan. We have measured the changes of the collimated X-ray beam size and intensity at different distances from the collimator. We have also studied the effects of beam size and intensity in the reconstruction of XLCT. PMID:27446686

  17. Multiple pinhole collimator based X-ray luminescence computed tomography

    PubMed Central

    Zhang, Wei; Zhu, Dianwen; Lun, Michael; Li, Changqing

    2016-01-01

    X-ray luminescence computed tomography (XLCT) is an emerging hybrid imaging modality, which is able to improve the spatial resolution of optical imaging to hundreds of micrometers for deep targets by using superfine X-ray pencil beams. However, due to the low X-ray photon utilization efficiency in a single pinhole collimator based XLCT, it takes a long time to acquire measurement data. Herein, we propose a multiple pinhole collimator based XLCT, in which multiple X-ray beams are generated to scan a sample at multiple positions simultaneously. Compared with the single pinhole based XLCT, the multiple X-ray beam scanning method requires much less measurement time. Numerical simulations and phantom experiments have been performed to demonstrate the feasibility of the multiple X-ray beam scanning method. In one numerical simulation, we used four X-ray beams to scan a cylindrical object with 6 deeply embedded targets. With measurements from 6 angular projections, all 6 targets have been reconstructed successfully. In the phantom experiment, we generated two X-ray pencil beams with a collimator manufactured in-house. Two capillary targets with 0.6 mm edge-to-edge distance embedded in a cylindrical phantom have been reconstructed successfully. With the two beam scanning, we reduced the data acquisition time by 50%. From the reconstructed XLCT images, we found that the Dice similarity of targets is 85.11% and the distance error between two targets is less than 3%. We have measured the radiation dose during XLCT scan and found that the radiation dose, 1.475 mSv, is in the range of a typical CT scan. We have measured the changes of the collimated X-ray beam size and intensity at different distances from the collimator. We have also studied the effects of beam size and intensity in the reconstruction of XLCT. PMID:27446686

  18. Intensity modulated neutron radiotherapy optimization by photon proxy

    SciTech Connect

    Snyder, Michael; Hammoud, Ahmad; Bossenberger, Todd; Spink, Robyn; Burmeister, Jay

    2012-08-15

    Purpose: Introducing intensity modulation into neutron radiotherapy (IMNRT) planning has the potential to mitigate some normal tissue complications seen in past neutron trials. While the hardware to deliver IMNRT plans has been in use for several years, until recently the IMNRT planning process has been cumbersome and of lower fidelity than conventional photon plans. Our in-house planning system used to calculate neutron therapy plans allows beam weight optimization of forward planned segments, but does not provide inverse optimization capabilities. Commercial treatment planning systems provide inverse optimization capabilities, but currently cannot model our neutron beam. Methods: We have developed a methodology and software suite to make use of the robust optimization in our commercial planning system while still using our in-house planning system to calculate final neutron dose distributions. Optimized multileaf collimator (MLC) leaf positions for segments designed in the commercial system using a 4 MV photon proxy beam are translated into static neutron ports that can be represented within our in-house treatment planning system. The true neutron dose distribution is calculated in the in-house system and then exported back through the MATLAB software into the commercial treatment planning system for evaluation. Results: The planning process produces optimized IMNRT plans that reduce dose to normal tissue structures as compared to 3D conformal plans using static MLC apertures. The process involves standard planning techniques using a commercially available treatment planning system, and is not significantly more complex than conventional IMRT planning. Using a photon proxy in a commercial optimization algorithm produces IMNRT plans that are more conformal than those previously designed at our center and take much less time to create. Conclusions: The planning process presented here allows for the optimization of IMNRT plans by a commercial treatment planning

  19. Collimator performance evaluation for In-111 SPECT using a detection/localization task

    NASA Astrophysics Data System (ADS)

    Lu, Yihuan; Chen, Lin; Gindi, Gene

    2014-02-01

    In SPECT, the collimator is a crucial element in controlling image quality. We take a task performance approach to collimator performance evaluation in which an ideal observer is applied to the raw camera data without regard to the subsequent reconstruction stage. The clinical context of our collimator study is one of searching for and detecting neuroendocrine tumor metastases in the liver as seen in In-111 Octreotide SPECT. Our task involves detection and localization of a signal and thus differs from the conventionally used detection-only task. The scalar task performance metric is ALROC, the area under the localization receiver operating characteristic curve. Since In-111 emits photons at both 171 and 245 keV, the higher energy emissions can contribute significant septal scatter and penetration. Our collimator evaluations address a question previously considered by Mähler et al (2012 IEEE Trans. Nucl. Sci. 59 47-53) who used a different methodology: does allowing a limited amount of septal scatter and penetration yield improved task performance? We used simulation methods to evaluate five parallel-hole collimators. The collimators had roughly equal geometric sensitivity and resolution but a range of contributions from septal effects leading to variations in total sensitivity and resolution. We found that the best performance was obtained with a collimator that allowed a moderate amount of septal scatter and penetration.

  20. Molecular imaging of small animals with a triple-head SPECT system using pinhole collimation.

    PubMed

    Metzler, S D; Jaszczak, R J; Patil, N H; Vemulapalli, S; Akabani, G; Chin, B B

    2005-07-01

    Pinhole collimation yields high sensitivity when the distance from the object to the aperture is small, as in the case of imaging small animals. Fine-resolution images may be obtained when the magnification is large since this mitigates the effect of detector resolution. Large magnifications in pinhole single-photon emission computed tomography (SPECT) may be obtained by using a collimator whose focal length is many times the radius of rotation. This may be achieved without truncation if the gamma camera is large. We describe a commercially available clinical scanner mated with pinhole collimation and an external linear stage. The pinhole collimation gives high magnification. The linear stage allows for helical pinhole SPECT. We have used the system to image radiolabeled molecules in phantoms and small animals.

  1. A 3 × 3 silicon drift chamber array for application in an electronic collimator

    NASA Astrophysics Data System (ADS)

    Kuykens, H. J. P.; Audet, S. A.

    1988-12-01

    The mechanical collimator, used in combination with an Anger camera in nuclear-medical imaging studies, can be replaced with a radiation detector, which performs the function of electronic collimation. The two-detector system is based on the Compton scattering of incident gamma-photons. Although germanium is more efficient in the absorption of gamma-photons by the Compton process than silicon, silicon has advantages which cannot be overlooked when choosing a substrate material on which to fabricate the radiation detector to be used in the electronic collimator system. The semiconductor drift chamber (SDC) principle offers advantages in the construction of a two-dimensional radiation detector array. Simulations of the potential field within the detector array will be shown, as well as the design and processing sequence used in the fabrication of a 3 × 3 SDC array.

  2. High-intensity beam collimation and targetry

    SciTech Connect

    Mokhov, N.V.; /Fermilab

    2006-11-01

    Principles, design criteria and realization of reliable collimation systems for the high-power accelerators and hadron colliders are described. Functionality of collimators as the key elements of the machine protection system are discussed along with the substantial progress on the crystal collimation front. The key issues are considered in design of high-power target systems and achieving their best performance. Simulation code requirements are presented.

  3. Recent Progress on the Design of a Rotatable Copper Collimator for the LHC Collimation Upgrade

    SciTech Connect

    Smith, Jeffrey Claiborne; Keller, Lewis; Lundgren, Steven; Markiewicz, Thomas Walter; Lari, Luisella; /Cern /EPFL-ISIC

    2009-08-03

    The Phase II upgrade to the LHC collimation system calls for complementing the 30 high robust Phase I graphite collimators with 30 high Z Phase II collimators. One option is to use metallic rotatable collimators and this design will be discussed here. The Phase II collimators must be robust in various operating conditions and accident scenarios. Design issues include: (1) Collimator jaw deflection due to heating and sagitta must be small when operated in the steady state condition, (2) Collimator jaws must withstand transitory periods of high beam impaction with no permanent damage, (3) Jaws must recover from accident scenario where up to 8 full intensity beam pulses impact on the jaw surface and (4) The beam impedance contribution due to the collimators must be small to minimize coherent beam instabilities. This paper reports on recent updates to the design and testing.

  4. Linac Coherent Light Source Electron Beam Collimation

    SciTech Connect

    Wu, J.; Dowell, D.; Emma, P.; Limborg-Deprey, C.; Schmerge, J.F.; /SLAC

    2007-04-27

    This paper describes the design and simulation of the electron beam collimation system in the Linac Coherent Light Source (LCLS). Dark current is expected from the gun and some of the accelerating cavities. Particle tracking of the expected dark current through the entire LCLS linac, from gun through FEL undulator, is used to estimate final particle extent in the undulator as well as expected beam loss at each collimator or aperture restriction. A table of collimators and aperture restrictions is listed along with halo particle loss results, which includes an estimate of average continuous beam power lost. In addition, the transverse wakefield alignment tolerances are calculated for each collimator.

  5. Beam halo collimation in heavy ion synchrotrons

    NASA Astrophysics Data System (ADS)

    Strašík, I.; Prokhorov, I.; Boine-Frankenheim, O.

    2015-08-01

    This paper presents a systematic study of the halo collimation of ion beams from proton up to uranium in synchrotrons. The projected Facility for Antiproton and Ion Research synchrotron SIS100 is used as a reference case. The concepts are separated into fully stripped (e.g., 238U92+ ) and partially stripped (e.g., 238U28+ ) ion collimation. An application of the two-stage betatron collimation system, well established for proton accelerators, is intended also for fully stripped ions. The two-stage system consists of a primary collimator (a scattering foil) and secondary collimators (bulky absorbers). Interaction of the particles with the primary collimator (scattering, momentum losses, and nuclear interactions) was simulated by using fluka. Particle-tracking simulations were performed by using mad-x. Finally, the dependence of the collimation efficiency on the primary ion species was determined. The influence of the collimation system adjustment, lattice imperfections, and beam parameters was estimated. The concept for the collimation of partially stripped ions employs a thin stripping foil in order to change their charge state. These ions are subsequently deflected towards a dump location using a beam optical element. The charge state distribution after the stripping foil was obtained from global. The ions were tracked by using mad-x.

  6. Design and performance evaluation of a 20-aperture multipinhole collimator for myocardial perfusion imaging applications

    NASA Astrophysics Data System (ADS)

    Bowen, Jason D.; Huang, Qiu; Ellin, Justin R.; Lee, Tzu-Cheng; Shrestha, Uttam; Gullberg, Grant T.; Seo, Youngho

    2013-10-01

    Single photon emission computed tomography (SPECT) myocardial perfusion imaging remains a critical tool in the diagnosis of coronary artery disease. However, after more than three decades of use, photon detection efficiency remains poor and unchanged. This is due to the continued reliance on parallel-hole collimators first introduced in 1964. These collimators possess poor geometric efficiency. Here we present the performance evaluation results of a newly designed multipinhole collimator with 20 pinhole apertures (PH20) for commercial SPECT systems. Computer simulations and numerical observer studies were used to assess the noise, bias and diagnostic imaging performance of a PH20 collimator in comparison with those of a low energy high resolution (LEHR) parallel-hole collimator. Ray-driven projector/backprojector pairs were used to model SPECT imaging acquisitions, including simulation of noiseless projection data and performing MLEM/OSEM image reconstructions. Poisson noise was added to noiseless projections for realistic projection data. Noise and bias performance were investigated for five mathematical cardiac and torso (MCAT) phantom anatomies imaged at two gantry orbit positions (19.5 and 25.0 cm). PH20 and LEHR images were reconstructed with 300 MLEM iterations and 30 OSEM iterations (ten subsets), respectively. Diagnostic imaging performance was assessed by a receiver operating characteristic (ROC) analysis performed on a single MCAT phantom; however, in this case PH20 images were reconstructed with 75 pixel-based OSEM iterations (four subsets). Four PH20 projection views from two positions of a dual-head camera acquisition and 60 LEHR projections were simulated for all studies. At uniformly-imposed resolution of 12.5 mm, significant improvements in SNR and diagnostic sensitivity (represented by the area under the ROC curve, or AUC) were realized when PH20 collimators are substituted for LEHR parallel-hole collimators. SNR improves by factors of 1

  7. Development of an efficient scanning and purging magnet system for IMRT with narrow high energy photon beams

    NASA Astrophysics Data System (ADS)

    Andreassen, Björn; Svensson, Roger; Holmberg, Rickard; Danared, Håkan; Brahme, Anders

    2009-12-01

    Due to the clinical advantages of Intensity Modulated Radiation Therapy (IMRT) high flexibility and accuracy in intensity modulated dose delivery is desirable to really maximize treatment outcome. Although it is possible to deliver IMRT by using broad beams in combination with dynamic multileaf collimation the process is rather time consuming and inefficient. By using narrow scanned high energy photon beams the treatment outcome can be improved, the treatment time reduced and accurate 3D in vivo dose delivery monitoring is possible by PET-CT based dose delivery imaging of photo nuclear reactions in human tissues. Narrow photon beams can be produced by directing a low emittance high energy electron beam on a thin target, and then cleaning the therapeutic photon beam from transmitted high energy electrons, and photon generated charged leptons, with a dedicated purging magnet placed directly downstream of the target. To have an effective scanning and purging magnet system the purging magnet should be placed immediately after the bremsstrahlung target to deflect the transmitted electrons to an efficient electron stopper. In the static electron stopper the electrons should be safely collected independent of the desired direction of the therapeutic scanned photon beam. The SID (Source to Isocenter Distance) should preferably be short while retaining the ability to scan over a large area on the patient and consequently there are severe requirements both on the strength and the geometry of the scanning and purging magnets. In the present study an efficient magnet configuration with a purging and scanning magnet assembly is developed for electron energies in the 50-75 MeV range and a SID of 75 cm. For a bremsstrahlung target of 3 mm Be these electron energies produce a photon beam of 25-17 mm FWHM (Full Width Half Maximum) at a SID of 75 cm. The magnet system was examined both in terms of the efficiency in scanning the narrow bremsstrahlung beam and the deflection of

  8. Ion source with improved primary arc collimation

    DOEpatents

    Dagenhart, W.K.

    1983-12-16

    An improved negative ion source is provided in which a self-biasing, molybdenum collimator is used to define the primary electron stream arc discharge from a filament operated at a negative potential. The collimator is located between the anode and the filament. It is electrically connected to the anode by means of an appropriate size resistor such that the collimator is biased at essentially the filament voltage during operation. Initially, the full arc voltage appears across the filament to collimator until the arc discharge strikes. Then the collimator biases itself to essentially filament potential due to current flow through the resistor thus defining the primary electron stream without intercepting any appreciable arc power. The collimator aperture is slightly smaller than the anode aperture to shield the anode from the arc power which, in the past, has caused overheating and erosion of the anode collimator during extended time pulsed-beam operation of the source. With the self-biasing collimator of this invention, the ion source may be operated from short pulse periods to steady-state without destroying the anode.

  9. Ion source with improved primary arc collimation

    DOEpatents

    Dagenhart, William K.

    1985-01-01

    An improved negative ion source is provided in which a self-biasing, molybdenum collimator is used to define the primary electron stream arc discharge from a filament operated at a negative potential. The collimator is located between the anode and the filament. It is electrically connected to the anode by means of an appropriate size resistor such that the collimator is biased at essentially the filament voltage during operation. Initially, the full arc voltage appears across the filament to collimator until the arc discharge strikes. Then the collimator biases itself to essentially filament potential due to current flow through the resistor thus defining the primary electron stream without intercepting any appreciable arc power. The collimator aperture is slightly smaller than the anode aperture to shield the anode from the arc power, thereby preventing the exposure of the anode to the full arc power which, in the past, has caused overheating and erosion of the anode collimator during extended time pulsed-beam operation of the source. With the self-biasing collimator of this invention, the ion source may be operated from short pulse periods to steady-state without destroying the anode.

  10. Dual-prism interferometer for collimation testing

    SciTech Connect

    Hii, King Ung; Kwek, Kuan Hiang

    2009-01-10

    An air-wedge lateral-shear interferometer using two prisms is presented. With a variable shear, the interferometer is suitable for testing collimation of a wide range of beam sizes down to a few millimeters in diameter. No antireflection coatings are necessary. Collimation for a light source with short coherent length is also demonstrated.

  11. Construction and Bench Testing of a Rotatable Collimator for the LHC Collimation Upgrade

    SciTech Connect

    Smith, Jeffrey Claiborne; Keller, Lewis; Lundgren, Steven; Markiewicz, Thomas; /SLAC

    2010-08-26

    The Phase II upgrade to the LHC collimation system calls for complementing the 30 high robust Phase I graphite secondary collimators with 30 high Z Phase II collimators. The Phase II collimators must be robust in various operating conditions and accident scenarios. This paper reports on the final construction and testing of the prototype collimator to be installed in the SPS (Super Proton Synchrotron) at CERN. Bench-top measurements will demonstrate that the device is fully operational and has the mechanical and vacuum characteristics acceptable for installation in the SPS.

  12. The PEP-II Movable Collimators

    SciTech Connect

    DeBarger, S.; Metcalfe, S.; Ng, C.; Porter, T.G.; Seeman, J.; Sullivan, M.; Wienands, U.; /SLAC

    2006-03-13

    Three movable collimators have been manufactured for installation in the PEP-II LER and HER beamlines upstream of BaBar to improve backgrounds in BaBar by a factor of 2. Each collimator has a pair of horizontally opposed, water cooled jaws with RF finger seals all around the edge of the jaws, these seals are the only sliding parts inside the vacuum chamber. Each jaw travels independently through a distance of 16.5 mm (LER) or 21mm (HER) and is supported above the collimator from motorized slideways with position feedback. The larger HER collimator has a titanium sublimation pump incorporated into the underside of the collimator, pumping through RF screens in the bottom of the chamber. Water cooled fixed ramps protect the leading and trailing edges of the jaws.

  13. Prototype Testing for a Copper Rotatable Collimator for the LHC Collimation Upgrade

    SciTech Connect

    Smith, Jeffrey Claiborne; Anzalone, Gene; Doyle, Eric; Keller, Lewis; Lundgren, Steven; Markiewicz, Thomas Walter; Rogers, Reggie; /SLAC

    2009-01-20

    The Phase II upgrade to the LHC collimation system calls for complementing the robust Phase I graphite collimators with high Z Phase II collimators. The design for the collimation upgrade has not been finalized. One option is to use metallic rotatable collimators and testing of this design will be discussed here. The Phase II collimators must be robust in various operating conditions and accident scenarios. A prototype collimator jaw referred to as RC0 has been tested for both mechanical and thermal compliance with the design goals. Thermal expansion bench-top tests are compared to ANSYS simulation results. The prototype has also been tested in vacuum bake-out to confirm compliance with the LHC vacuum spec. CMM equipment has been used to verify the flatness of the jaw surface after heat tests and bake-out.

  14. Soller collimators for small angle neutron scattering

    SciTech Connect

    Crawford, R.K.; Epperson, J.E.; Thiyagarajan, P.

    1988-09-30

    Small angle diffractometers at pulsed sources need to have fairly short flight paths if they are to make use of the long-wavelength portion of the spectrum without encountering problems from frame overlap or sacrificing intensity with band-limiting or pulse-removing choppers. With such short flight paths, achieving the necessary angular collimation in the incident beam while utilizing the full source size (/approximately/10 cm diameter) and a reasonable sample size (/approximately/1 cm diameter) requires the use of converging multiple-aperture collimation. If the collimation channels are all focused to the same point on the detector then the large sample size will not affect Q/sub min/ or the Q-resolution, even if the sample-to-detector distance is short. The Small Angle Diffractometer (SAD) at IPNS uses crossed converging soller collimators to provide focusing multiple-aperture collimation having /approximately/400 converging beam channels with essentially no ''dead'' space between them. This entire collimator system occupies a distance of only /approximately/60 cm along the incident flight path, while providing angular collimation of 0.003 radians FWHM. The dimensions for the SAD upstream collimator are L/sub c/ = 32.8 cm, d/sub 1/ = 0.974 mm, d/sub 2/ = 0.851 mm, while for the SAD downstream collimator L/sub c/ = 25.0 cm, d/sub 1/ - 0.844 mm, d/sub 2/ = 0.750 mm. Each of these collimators has 20 blades defining 21 horizontal or vertical channels. 4 refs., 6 figs.

  15. Optical cone beam tomography of Cherenkov-mediated signals for fast 3D dosimetry of x-ray photon beams in water

    SciTech Connect

    Glaser, Adam K. E-mail: Brian.W.Pogue@dartmouth.edu; Andreozzi, Jacqueline M.; Zhang, Rongxiao; Pogue, Brian W. E-mail: Brian.W.Pogue@dartmouth.edu; Gladstone, David J.

    2015-07-15

    Purpose: To test the use of a three-dimensional (3D) optical cone beam computed tomography reconstruction algorithm, for estimation of the imparted 3D dose distribution from megavoltage photon beams in a water tank for quality assurance, by imaging the induced Cherenkov-excited fluorescence (CEF). Methods: An intensified charge-coupled device coupled to a standard nontelecentric camera lens was used to tomographically acquire two-dimensional (2D) projection images of CEF from a complex multileaf collimator (MLC) shaped 6 MV linear accelerator x-ray photon beam operating at a dose rate of 600 MU/min. The resulting projections were used to reconstruct the 3D CEF light distribution, a potential surrogate of imparted dose, using a Feldkamp–Davis–Kress cone beam back reconstruction algorithm. Finally, the reconstructed light distributions were compared to the expected dose values from one-dimensional diode scans, 2D film measurements, and the 3D distribution generated from the clinical Varian ECLIPSE treatment planning system using a gamma index analysis. A Monte Carlo derived correction was applied to the Cherenkov reconstructions to account for beam hardening artifacts. Results: 3D light volumes were successfully reconstructed over a 400 × 400 × 350 mm{sup 3} volume at a resolution of 1 mm. The Cherenkov reconstructions showed agreement with all comparative methods and were also able to recover both inter- and intra-MLC leaf leakage. Based upon a 3%/3 mm criterion, the experimental Cherenkov light measurements showed an 83%–99% pass fraction depending on the chosen threshold dose. Conclusions: The results from this study demonstrate the use of optical cone beam computed tomography using CEF for the profiling of the imparted dose distribution from large area megavoltage photon beams in water.

  16. Optical cone beam tomography of Cherenkov-mediated signals for fast 3D dosimetry of x-ray photon beams in water

    PubMed Central

    Glaser, Adam K.; Andreozzi, Jacqueline M.; Zhang, Rongxiao; Pogue, Brian W.; Gladstone, David J.

    2015-01-01

    Purpose: To test the use of a three-dimensional (3D) optical cone beam computed tomography reconstruction algorithm, for estimation of the imparted 3D dose distribution from megavoltage photon beams in a water tank for quality assurance, by imaging the induced Cherenkov-excited fluorescence (CEF). Methods: An intensified charge-coupled device coupled to a standard nontelecentric camera lens was used to tomographically acquire two-dimensional (2D) projection images of CEF from a complex multileaf collimator (MLC) shaped 6 MV linear accelerator x-ray photon beam operating at a dose rate of 600 MU/min. The resulting projections were used to reconstruct the 3D CEF light distribution, a potential surrogate of imparted dose, using a Feldkamp–Davis–Kress cone beam back reconstruction algorithm. Finally, the reconstructed light distributions were compared to the expected dose values from one-dimensional diode scans, 2D film measurements, and the 3D distribution generated from the clinical Varian ECLIPSE treatment planning system using a gamma index analysis. A Monte Carlo derived correction was applied to the Cherenkov reconstructions to account for beam hardening artifacts. Results: 3D light volumes were successfully reconstructed over a 400 × 400 × 350 mm3 volume at a resolution of 1 mm. The Cherenkov reconstructions showed agreement with all comparative methods and were also able to recover both inter- and intra-MLC leaf leakage. Based upon a 3%/3 mm criterion, the experimental Cherenkov light measurements showed an 83%–99% pass fraction depending on the chosen threshold dose. Conclusions: The results from this study demonstrate the use of optical cone beam computed tomography using CEF for the profiling of the imparted dose distribution from large area megavoltage photon beams in water. PMID:26133613

  17. Collimation of average multiplicity in QCD jets

    NASA Astrophysics Data System (ADS)

    Arleo, François; Pérez Ramos, Redamy

    2009-11-01

    The collimation of average multiplicity inside quark and gluon jets is investigated in perturbative QCD in the modified leading logarithmic approximation (MLLA). The role of higher order corrections accounting for energy conservation and the running of the coupling constant leads to smaller multiplicity collimation as compared to leading logarithmic approximation (LLA) results. The collimation of jets produced in heavy-ion collisions has also been explored by using medium-modified splitting functions enhanced in the infrared sector. As compared to elementary collisions, the angular distribution of the jet multiplicity is found to broaden in QCD media at all energy scales.

  18. Geometrical Wake of a Smooth Flat Collimator

    SciTech Connect

    Stupakov, G.V.; /SLAC

    2011-09-09

    A transverse geometrical wake generated by a beam passing through a smooth flat collimator with a gradually varying gap between the upper and lower walls is considered. Based on generalization of the approach recently developed for a smooth circular taper we reduce the electromagnetic problem of the impedance calculation to the solution of two much simpler static problems - a magnetostatic and an electrostatic ones. The solution shows that in the limit of not very large frequencies, the impedance increases with the ratio h/d where h is the width and d is the distance between the collimating jaws. Numerical results are presented for the NLC Post Linac collimator.

  19. Design Studies of a CZT-based Detector Combined with a Pixel-Geometry-Matching Collimator for SPECT Imaging

    PubMed Central

    Weng, Fenghua; Bagchi, Srijeeta; Huang, Qiu; Seo, Youngho

    2014-01-01

    Single Photon Emission Computed Tomography (SPECT) suffers limited efficiency due to the need for collimators. Collimator properties largely decide the data statistics and image quality. Various materials and configurations of collimators have been investigated in many years. The main thrust of our study is to evaluate the design of pixel-geometry-matching collimators to investigate their potential performances using Geant4 Monte Carlo simulations. Here, a pixel-geometry-matching collimator is defined as a collimator which is divided into the same number of pixels as the detector’s and the center of each pixel in the collimator is a one-to-one correspondence to that in the detector. The detector is made of Cadmium Zinc Telluride (CZT), which is one of the most promising materials for applications to detect hard X-rays and γ-rays due to its ability to obtain good energy resolution and high light output at room temperature. For our current project, we have designed a large-area, CZT-based gamma camera (20.192 cm×20.192 cm) with a small pixel pitch (1.60 mm). The detector is pixelated and hence the intrinsic resolution can be as small as the size of the pixel. Materials of collimator, collimator hole geometry, detection efficiency, and spatial resolution of the CZT detector combined with the pixel-matching collimator were calculated and analyzed under different conditions. From the simulation studies, we found that such a camera using rectangular holes has promising imaging characteristics in terms of spatial resolution, detection efficiency, and energy resolution. PMID:25378898

  20. Acoustic logic gates and Boolean operation based on self-collimating acoustic beams

    SciTech Connect

    Zhang, Ting; Xu, Jian-yi; Cheng, Ying Liu, Xiao-jun; Guo, Jian-zhong

    2015-03-16

    The reveal of self-collimation effect in two-dimensional (2D) photonic or acoustic crystals has opened up possibilities for signal manipulation. In this paper, we have proposed acoustic logic gates based on the linear interference of self-collimated beams in 2D sonic crystals (SCs) with line-defects. The line defects on the diagonal of the 2D square SCs are actually functioning as a 3 dB splitter. By adjusting the phase difference between two input signals, the basic Boolean logic functions such as XOR, OR, AND, and NOT are achieved both theoretically and experimentally. Due to the non-diffracting property of self-collimation beams, more complex Boolean logic and algorithms such as NAND, NOR, and XNOR can be realized by cascading the basic logic gates. The achievement of acoustic logic gates and Boolean operation provides a promising approach for acoustic signal computing and manipulations.

  1. Collimation of laser-produced proton beam

    NASA Astrophysics Data System (ADS)

    Takano, M.; Nagashima, T.; Izumiyama, T.; Gu, Y. J.; Barada, D.; Kong, Q.; Wang, P. X.; Ma, Y. Y.; Wang, W. M.; Kawata, S.

    2016-03-01

    In intense laser plasma interaction for particle acceleration several issues remain to be solved. In this paper we focus on a collimation of ion beam, which is produced by a laser plasma interaction. In this study, the ion beam is collimated by a thin film target. When an intense short pulse laser illuminates a target, target electrons are accelerated, and create an electron cloud that generates a sheath electric field at the target surface. Such the ion acceleration mechanism is called the target normal sheath acceleration (TNSA). The TNSA field would be used for the ion beam collimation by the electric field. We have successfully obtained a collimated beam in our particle-in-cell simulations.

  2. Variable aperture collimator for high energy radiation

    DOEpatents

    Hill, Ronald A.

    1984-05-22

    An apparatus is disclosed providing a variable aperture energy beam collimator. A plurality of beam opaque blocks are in sliding interface edge contact to form a variable aperture. The blocks may be offset at the apex angle to provide a non-equilateral aperture. A plurality of collimator block assemblies may be employed for providing a channel defining a collimated beam. Adjacent assemblies are inverted front-to-back with respect to one another for preventing noncollimated energy from emerging from the apparatus. An adjustment mechanism comprises a cable attached to at least one block and a hand wheel mechanism for operating the cable. The blocks are supported by guide rods engaging slide brackets on the blocks. The guide rods are pivotally connected at each end to intermediate actuators supported on rotatable shafts to change the shape of the aperture. A divergent collimated beam may be obtained by adjusting the apertures of adjacent stages to be unequal.

  3. Solid Collection Efforts: Ta Collimator Evaluation

    SciTech Connect

    Gostic, J M

    2011-11-21

    Ta collimator sets that were part of the gated x-ray detector diagnostic (GXD) at NIF were analyzed for debris distribution and damage in 2011. These disks (ranging in thickness from 250 to 750 {mu}m) were fielded approximately 10 cm from target chamber center (TCC) on various symcap, THD and re-emit shots. The nose cone holder and forward Ta collimator (facing target chamber center, TCC) from all shots show evidence of surface melt. Non-destructive analysis techniques such as optical microscopy, surface profilometry, scanning electron microscopy (SEM), energy dispersive spectroscopy (EDS), and x-ray fluorescence (XRF) were used to determine debris composition and degree of deformation associated with each Ta disk. Molten debris from the stainless steel nose cone contaminated the surface of the collimators along with other debris associated with the target assembly (Al, Si, Cu, Au and In). Surface elemental analysis of the forward collimator Ta disks indicates that Au hohlraum debris is less concentrated on these samples versus those fielded 50 cm from TCC in the wedge range filter (WRF) assembly. It is possible that the Au is distributed below or within the stainless steel melt layer covering the disk, as most of the foreign debris is captured in the melted coating. The other disks (fielded directly behind the forward collimator in a sandwiched configuration) have visible forms of deformation and warping. The degree of warping increases as the shock wave penetrates the assembly with the most damage sustained on the back collimator. In terms of developing a solid collection capability, the collimator analyses suggests that close proximity may cause more interference with capsule debris collection and more damage to the surface of the collector diagnostic. The analyses of the Ta collimators were presented to the Target and Laser Interaction Sphere (TaLIS) group; a representative presentation is attached to this document.

  4. Comparison of Carbon and Hi-Z Primary Collimators for the LHC Phase II Collimation System

    SciTech Connect

    Keller, Lewis; Markiewicz, Thomas; Smith, Jeffrey; Assmann, Ralph; Bracco, Chiara; Weiler, Thomas; /Karlsruhe, Inst. Technol.

    2011-10-31

    A current issue with the LHC collimation system is single-diffractive, off-energy protons from the primary collimators that pass completely through the secondary collimation system and are absorbed immediately downbeam in the cold magnets of the dispersion suppressor section. Simulations suggest that the high impact rate could result in quenching of these magnets. We have studied replacing the 60 cm primary graphite collimators, which remove halo mainly by inelastic strong interactions, with 5.25 mm tungsten, which remove halo mainly by multiple coulomb scattering and thereby reduce the rate of single-diffractive interactions that cause losses in the dispersion suppressor.

  5. New Methods of Particle Collimation in Colliders

    SciTech Connect

    Stancari, Giulio; /Fermilab

    2011-10-01

    The collimation system is an essential part of the design of any high-power accelerator. Its functions include protection of components from accidental and intentional energy deposition, reduction of backgrounds, and beam diagnostics. Conventional multi-stage systems based on scatterers and absorbers offer robust shielding and efficient collection of losses. Two complementary concepts have been proposed to address some of the limitations of conventional systems: channeling and volume reflection in bent crystals and collimation with hollow electron beams. The main focus of this paper is the hollow electron beam collimator, a novel concept based on the interaction of the circulating beam with a 5-keV, magnetically confined, pulsed hollow electron beam in a 2-m-long section of the ring. The electrons enclose the circulating beam, kicking halo particles transversely and leaving the beam core unperturbed. By acting as a tunable diffusion enhancer and not as a hard aperture limitation, the hollow electron beam collimator extends conventional collimation systems beyond the intensity limits imposed by tolerable losses. The concept was tested experimentally at the Fermilab Tevatron proton-antiproton collider. Results on the collimation of 980-GeV antiprotons are presented, together with prospects for the future.

  6. Wake fields in SLAC Linac Collimators

    SciTech Connect

    Novokhatski, Alexander; Decker, F. -J.; Smith, H.; Sullivan, M.

    2014-12-02

    When a beam travels near collimator jaws, it gets an energy loss and a transverse kick due to the backreaction of the beam field diffracted from the jaws. The effect becomes very important for an intense short bunch when a tight collimation of the background beam halo is required. In the Linac Coherent Light Source at SLAC a collimation system is used to protect the undulators from radiation due to particles in the beam halo. The halo is most likely formed from gun dark current or dark current in some of the accelerating sections. However, collimators are also responsible for the generation of wake fields. The wake field effect from the collimators not only brings an additional energy jitter and change in the trajectory of the beam, but it also rotates the beam on the phase plane, which consequently leads to a degradation of the performance of the Free Electron Laser at the Linac Coherent Light Source. In this paper, we describe a model of the wake field radiation in the SLAC linac collimators. We use the results of a numerical simulation to illustrate the model. Based on the model, we derive simple formulas for the bunch energy loss and the average kick. In addition, we also present results from experimental measurements that confirm our model.

  7. Beam characteristics of fiber-based supercontinuum light sources with mirror- and lens-based beam collimators.

    PubMed

    Arnold, Ian J; Moosmüller, Hans; Sharma, Noopur; Mazzoleni, Claudio

    2014-06-01

    Commercially available supercontinuum light sources that cover most of the solar spectrum are well suited for instrumentation, where a well-collimated beam with wide spectral coverage is needed. Typically, the optical power is emitted from a single-mode photonic-crystal fiber and the output can either be collimated using a proprietary, permanently integrated, lens-based collimator or with a customer-provided, off-axis parabolic mirror. Here, we evaluate both approaches and conclude that, superior beam quality and collimation over the whole spectral range can be obtained with an off-axis parabolic mirror, however at the price of a more complex and bulky system requiring additional user alignment. PMID:24921577

  8. Upgrade scenario for the RHIC collimation system

    SciTech Connect

    Robert-Demolaize, G.; Drees, A.

    2012-01-19

    The RHIC collimation system is used to reduce background levels in both STAR and PHENIX detectors. With a push for higher luminosity in the near future, it becomes critical to check if and how the level of performance of the collimators can be improved. The following reviews a proposal for additional collimators placed further downstream of the current system and designed to intercept the tertiary halo coming out of the IR8 insertion before it can reach the triplet quadrupoles in either STAR or PHENIX. Simulations have been peformed to quantify the efficiency of additional collimator jaws in RHIC. Each figure presented in this article clearly shows that the additional mask collimators provide the expected reduction in losses around the machine, and especially to the incoming triplet to the STAR experiment (IP6), for the Yellow beam as much as for the Blue beam. Looking at compiled statistics for all three working point cases studied, proton losses around the machine are reduced by roughly one order of magnitude: at most a factor 30 for magnet losses, and at most a factor 40 for losses in spaces between magnets.

  9. Collimator design for experimental minibeam radiation therapy

    SciTech Connect

    Babcock, Kerry; Sidhu, Narinder; Kundapur, Vijayananda; Ali, Kaiser

    2011-04-15

    Purpose: To design and optimize a minibeam collimator for minibeam radiation therapy studies using a 250 kVp x-ray machine as a simulated synchrotron source. Methods: A Philips RT250 orthovoltage x-ray machine was modeled using the EGSnrc/BEAMnrc Monte Carlo software. The resulting machine model was coupled to a model of a minibeam collimator with a beam aperture of 1 mm. Interaperture spacing and collimator thickness were varied to produce a minibeam with the desired peak-to-valley ratio. Results: Proper design of a minibeam collimator with Monte Carlo methods requires detailed knowledge of the x-ray source setup. For a cathode-ray tube source, the beam spot size, target angle, and source shielding all determine the final valley-to-peak dose ratio. Conclusions: A minibeam collimator setup was created, which can deliver a 30 Gy peak dose minibeam radiation therapy treatment at depths less than 1 cm with a valley-to-peak dose ratio on the order of 23%.

  10. A dosimetric comparison of real-time adaptive and non-adaptive radiotherapy: A multi-institutional study encompassing robotic, gimbaled, multileaf collimator and couch tracking

    PubMed Central

    Colvill, Emma; Booth, Jeremy; Nill, Simeon; Fast, Martin; Bedford, James; Oelfke, Uwe; Nakamura, Mitsuhiro; Poulsen, Per; Worm, Esben; Hansen, Rune; Ravkilde, Thomas; Scherman Rydhög, Jonas; Pommer, Tobias; Munck af Rosenschold, Per; Lang, Stephanie; Guckenberger, Matthias; Groh, Christian; Herrmann, Christian; Verellen, Dirk; Poels, Kenneth; Wang, Lei; Hadsell, Michael; Sothmann, Thilo; Blanck, Oliver; Keall, Paul

    2016-01-01

    Purpose A study of real-time adaptive radiotherapy systems was performed to test the hypothesis that, across delivery systems and institutions, the dosimetric accuracy is improved with adaptive treatments over non-adaptive radiotherapy in the presence of patient-measured tumor motion. Methods and materials Ten institutions with robotic(2), gimbaled(2), MLC(4) or couch tracking(2) used common materials including CT and structure sets, motion traces and planning protocols to create a lung and a prostate plan. For each motion trace, the plan was delivered twice to a moving dosimeter; with and without real-time adaptation. Each measurement was compared to a static measurement and the percentage of failed points for γ-tests recorded. Results For all lung traces all measurement sets show improved dose accuracy with a mean 2%/2 mm γ-fail rate of 1.6% with adaptation and 15.2% without adaptation (p < 0.001). For all prostate the mean 2%/2 mm γ-fail rate was 1.4% with adaptation and 17.3% without adaptation (p < 0.001). The difference between the four systems was small with an average 2%/2 mm γ-fail rate of <3% for all systems with adaptation for lung and prostate. Conclusions The investigated systems all accounted for realistic tumor motion accurately and performed to a similar high standard, with real-time adaptation significantly outperforming non-adaptive delivery methods. PMID:27016171

  11. Monte Carlo modeling of HD120 multileaf collimator on Varian TrueBeam linear accelerator for verification of 6X and 6X FFF VMAT SABR treatment plans.

    PubMed

    Bergman, Alanah M; Gete, Ermias; Duzenli, Cheryl; Teke, Tony

    2014-01-01

    A Monte Carlo (MC) validation of the vendor-supplied Varian TrueBeam 6 MV flattened (6X) phase-space file and the first implementation of the Siebers-Keall MC MLC model as applied to the HD120 MLC (for 6X flat and 6X flattening filter-free (6X FFF) beams) are described. The MC model is validated in the context of VMAT patient-specific quality assurance. The Monte Carlo commissioning process involves: 1) validating the calculated open-field percentage depth doses (PDDs), profiles, and output factors (OF), 2) adapting the Siebers-Keall MLC model to match the new HD120-MLC geometry and material composition, 3) determining the absolute dose conversion factor for the MC calculation, and 4) validating this entire linac/MLC in the context of dose calculation verification for clinical VMAT plans. MC PDDs for the 6X beams agree with the measured data to within 2.0% for field sizes ranging from 2 × 2 to 40 × 40 cm2. Measured and MC profiles show agreement in the 50% field width and the 80%-20% penumbra region to within 1.3 mm for all square field sizes. MC OFs for the 2 to 40 cm2 square fields agree with measurement to within 1.6%. Verification of VMAT SABR lung, liver, and vertebra plans demonstrate that measured and MC ion chamber doses agree within 0.6% for the 6X beam and within 2.0% for the 6X FFF beam. A 3D gamma factor analysis demonstrates that for the 6X beam, > 99% of voxels meet the pass criteria (3%/3 mm). For the 6X FFF beam, > 94% of voxels meet this criteria. The TrueBeam accelerator delivering 6X and 6X FFF beams with the HD120 MLC can be modeled in Monte Carlo to provide an independent 3D dose calculation for clinical VMAT plans. This quality assurance tool has been used clinically to verify over 140 6X and 16 6X FFF TrueBeam treatment plans.

  12. SU-E-T-534: Dosimetric Effect of Multileaf Collimator Leaf Width On Volumetric Modulated Arc Stereotactic Radiotherapy for Spine Tumors

    SciTech Connect

    Amoush, A; Djemil, T; Subedi, L; Huang, L; Xia, P

    2014-06-01

    Purpose: To study the dosimetric impact of MLC leaf width in patients treated with Volumetric Modulated Arc Therapy (VMAT) for spine Stereotactic Body radiation Therapy (SBRT). Methods: Twelve spine SBRT patients were retrospectively selected for this study. The patients were treated with IMRT following the RTOG-0631 of spine metastasis. The prescription dose was 16 Gy in one fraction to 90% of the target volume (V16 > 90%). The maximum spinal cord dose of 14 Gy and 10% of the cord receiving < 10 Gy (V10) were set as dose constraints. For purpose of this study, three dual arc VMAT plans were created for each patient using three different MLC leaf widths: 2.5 mm, 4mm, and 5mm. The compliance to RTOG 0631, conformal index (CI), dose gradient index (DGI), and number of monitor units (MUs) were compared. Results: The average V16 of the target was 91.91±1.36%, 93.73±2.38%, and 92.25±2.49% for 2.5 mm, 4 mm, and 5 mm leaf widths, respectively (p=0.39). Accordingly, the average CI was 1.36±0.39, 1.36±0.34, and 1.41±0.3 (0.96), respectively. The average DGI was 0.24 ± 0.05, 0.22 ± 0.05, and 0.23 ± 0.04, respectively (p=0.86). The average spinal cord maximum dose was 12.10 ± 0.88 Gy, 12.52 ± 1.15 Gy, and 12.05 ± 1.12 (p=0.75) and V10 was 2.69 ± 1.71 cc, 5.43 ± 2.16 cc, and 3.71 ± 2.34 cc (p=0.15) for 2.5 mm, 4 mm, and 5 mm leaf widths, respectively. According, the average number of MUs was 4255 ± 431 MU, 5049 ± 1036 MU, and 4231 ± 580 MU respectively (p=0.17). Conclusion: The use of 2.5 mm, 4 mm, and 5 mm MLCs achieved similar VMAT plan quality as recommended by RTOG-0631. The dosimetric parameters were also comparable for the three MLCs.

  13. Performance of a novel collimator for high-sensitivity brain SPECT

    SciTech Connect

    El Fakhri, Georges; Ouyang Jinsong; Zimmerman, Robert E.; Fischman, Alan J.; Kijewski, Marie Foley

    2006-01-15

    We assessed improvements in performance in detection and estimation tasks due to a novel brain single photon computed tomography collimator. Data were acquired on the CeraSPECT{sup TM} scanner using both new and standard collimators. The new variable focusing collimator SensOgrade{sup TM} samples the projections unequally, with central regions more heavily represented, to compensate for attenuation of counts from central brain structures. Furthermore, it utilizes more of the cylindrical crystal surface. Two phantom studies were performed. The first phantom was a 21-cm-diameter cylindrical background containing nine spheres ranging from 0.5 to 5 cm{sup 3} in volume. {sup 99m}Tc sphere to background activity ratio was 10:1. Twenty-nine 10-min datasets were acquired with each collimator. The second phantom was the Radiology Support Devices (Long Beach, CA) striatal phantom with striatal-background ratios of 10:1 on the left and 5:1 on the right. Twenty-nine 4-min datasets were acquired with each collimator. Perfusion imaging using {sup 99m}Tc-HMPAO was also performed in three healthy volunteers using both collimators under identical simulations. Projections were reconstructed by filtered backprojection with an unwindowed ramp filter. The nonprewhitening matched filter signal-to-noise ratio (NPW-SNR) was computed as a surrogate for human performance in detecting spherical lesions. Sphere activity concentration, radius, and location coordinates were simultaneously estimated by fitting images to an assumed model using an iterative nonlinear algorithm. Resolution recovery was implicit in the estimation procedure, as the point spread function was incorporated into the model. NPW-SNR for sphere detection was 1.5 to 2 times greater with the new collimator; for the striatal phantom the improvement in SNR was 54%. The SNR for estimating sphere activity concentration improved by 46 to 89 % for spheres located more than 5 cm from the phantom center. Images acquired with the

  14. Fermilab booster beam collimation and shielding

    SciTech Connect

    Nikolai V. Mokhov et al.

    2003-05-28

    The beam power in the upgraded Booster at 8 GeV and 10 Hz will be 64 kW. Beam loss can result in high radiation loads in the ring. The purpose of a new beam halo cleaning system is to localize proton losses in specially shielded regions. Calculations show that this 2-stage collimation system will localize about 99% of beam loss in straight sections 6 and 7 and immediately downstream. Beam loss in the rest of the machine will be on average 0.1W/m. Local shielding will provide tolerable prompt and residual radiation levels in the tunnel, above the tunnel at the surface and in the sump water. Results of thorough MARS calculations are presented for a new design which includes shielding integrated with the collimators, motors and controls ensuring a high performance and facilitating maintenance. First measurements of the collimation efficiency are presented.

  15. Collimation Studies with Hollow Electron Beams

    SciTech Connect

    Stancari, G.; Annala, G.; Johnson, T.R.; Saewert, G.W.; Shiltsev, V.; Still, D.A.; Valishev, A.; /Fermilab

    2011-08-01

    Recent experimental studies at the Fermilab Tevatron collider have shown that magnetically confined hollow electron beams can act as a new kind of collimator for high-intensity beams in storage rings. In a hollow electron beam collimator, electrons enclose the circulating beam. Their electric charge kicks halo particles transversely. If their distribution is axially symmetric, the beam core is unaffected. This device is complementary to conventional two-stage collimation systems: the electron beam can be placed arbitrarily close to the circulating beam; and particle removal is smooth, so that the device is a diffusion enhancer rather than a hard aperture limitation. The concept was tested in the Tevatron collider using a hollow electron gun installed in one of the existing electron lenses. We describe some of the technical aspects of hollow-beam scraping and the results of recent measurements.

  16. Laser beam collimation using Talbot interferometry

    NASA Technical Reports Server (NTRS)

    Ganesan, A. R.; Venkateswarlu, Putcha

    1993-01-01

    A modified method of checking laser beam collimation using a single grating and a right-angled prism is presented. The self-images (Talbot images) of a grating illuminated by a collimated beam are formed at some distance from the grating. The use of a right-angled prism makes it possible to carry out the folding of the self-image with respect to the original grating and to ensure that the grating lines in the self-image and the actual grating are inclined at equal angles with respect to horizontal direction. It is concluded that the proposed collimation test method has an in-built reference and does not require precise orientation of the grating as in the two-grating method. Large beams can be tested with a small-size assembly.

  17. Nonlinear resonant collimation for future linear colliders

    NASA Astrophysics Data System (ADS)

    Emma, P.; Helm, R.; Nosochkov, Y.; Pitthan, R.; Raubenheimer, T.; Thompson, K.; Zimmermann, F.

    1999-04-01

    We present a scheme for collimating large amplitude particles in the main linacs of a linear collider, by adding octupoles to the FODO lattice of the linac. With this scheme the requirements on downstream collimation can be greatly reduced or perhaps even eliminated. An analytic estimate of the amplitude at which particles are lost is made by calculating the separatrix of the fourth order resonance, and is in good agreement with the results of simulations. Simulations of particle distributions in the beam core and halo are presented, as well as alignment tolerances for the octupoles.

  18. Channeling collimation studies at the Fermilab Tevatron

    SciTech Connect

    Carrigan, Richard A.; Drozhdin, Alexandr I.; Fliller, Raymond P., III; Mokhov, Nikolai V.; Shiltsev, Vladimir D.; Still, Dean A.; /Fermilab

    2006-08-01

    Bent crystal channeling has promising advantages for accelerator beam collimation at high energy hadron facilities such as the LHC. This significance has been amplified by several surprising developments including multi-pass channeling and the observation of enhanced deflections over the entire arc of a bent crystal. The second effect has been observed both at RHIC and recently at the Tevatron. Results are reported showing channeling collimation of the circulating proton beam halo at the Tevatron. Parenthetically, this study is the highest energy proton channeling experiment ever carried out. The study is continuing.

  19. Verification measurements and clinical evaluation of the iPlan RT Monte Carlo dose algorithm for 6 MV photon energy

    NASA Astrophysics Data System (ADS)

    Petoukhova, A. L.; van Wingerden, K.; Wiggenraad, R. G. J.; van de Vaart, P. J. M.; van Egmond, J.; Franken, E. M.; van Santvoort, J. P. C.

    2010-08-01

    This study presents data for verification of the iPlan RT Monte Carlo (MC) dose algorithm (BrainLAB, Feldkirchen, Germany). MC calculations were compared with pencil beam (PB) calculations and verification measurements in phantoms with lung-equivalent material, air cavities or bone-equivalent material to mimic head and neck and thorax and in an Alderson anthropomorphic phantom. Dosimetric accuracy of MC for the micro-multileaf collimator (MLC) simulation was tested in a homogeneous phantom. All measurements were performed using an ionization chamber and Kodak EDR2 films with Novalis 6 MV photon beams. Dose distributions measured with film and calculated with MC in the homogeneous phantom are in excellent agreement for oval, C and squiggle-shaped fields and for a clinical IMRT plan. For a field with completely closed MLC, MC is much closer to the experimental result than the PB calculations. For fields larger than the dimensions of the inhomogeneities the MC calculations show excellent agreement (within 3%/1 mm) with the experimental data. MC calculations in the anthropomorphic phantom show good agreement with measurements for conformal beam plans and reasonable agreement for dynamic conformal arc and IMRT plans. For 6 head and neck and 15 lung patients a comparison of the MC plan with the PB plan was performed. Our results demonstrate that MC is able to accurately predict the dose in the presence of inhomogeneities typical for head and neck and thorax regions with reasonable calculation times (5-20 min). Lateral electron transport was well reproduced in MC calculations. We are planning to implement MC calculations for head and neck and lung cancer patients.

  20. The acceleration and collimation of jets.

    PubMed Central

    Begelman, M C

    1995-01-01

    I will discuss several issues related to the acceleration, collimation, and propagation of jets from active galactic nuclei. Hydromagnetic stresses provide the best bet for both accelerating relativistic flows and providing a certain amount of initial collimation. However, there are limits to how much "self-collimation" can be achieved without the help of an external pressurized medium. Moreover, existing models, which postulate highly organized poloidal flux near the base of the flow, are probably unrealistic. Instead, a large fraction of the magnetic energy may reside in highly disorganized "chaotic" fields. Such a field can also accelerate the flow to relativistic speeds, in some cases with greater efficiency than highly organized fields, but at the expense of self-collimation. The observational interpretation of jet physics is still hampered by a dearth of unambiguous diagnostics. Propagating disturbances in flows, such as the oblique shocks that may constitute the kiloparsec-scale "knots" in the M87 jet, may provide a wide range of untapped diagnostics for jet properties. PMID:11607615

  1. Modulation Collimator Fine Grids for HESP

    NASA Technical Reports Server (NTRS)

    Hecht, M.; Ling, J.; Siegel, P.; Kenny, T.; Grunthaner, F.

    1993-01-01

    The proposed High Engergy-Solar Physics (HESP) million will require fine, precise, high aspect ratio modulation collimators for the High Energy Imagin Spectrometer (HEISPEC). The fabrication approach described here rests on the proven precision of semiconductor lithography - the ability to replicate extremely accurate patterns onto different planar substrateds.

  2. Multiple wavelength light collimator and monitor

    NASA Technical Reports Server (NTRS)

    Gore, Warren J. (Inventor)

    2011-01-01

    An optical system for receiving and collimating light and for transporting and processing light received in each of N wavelength ranges, including near-ultraviolet, visible, near-infrared and mid-infrared wavelengths, to determine a fraction of light received, and associated dark current, in each wavelength range in each of a sequence of time intervals.

  3. BPM Design and Impedance Considerations for a Rotatable Collimator for the LHC Collimation Upgrade

    SciTech Connect

    Smith, Jeffrey Claiborne; Keller, Lewis; Lundgren, Steven; Markiewicz, Thomas; Young, Andrew; /SLAC

    2010-08-26

    The Phase II upgrade to the LHC collimation system calls for complementing the 30 high robust Phase I graphite secondary collimators with 30 high Z Phase II collimators. This paper reports on BPM and impedance considerations and measurements of the integrated BPMs in the prototype rotatable collimator to be installed in the Super Proton Synchrotron (SPS) at CERN. The BPMs are necessary to align the jaws with the beam. Without careful design the beam impedance can result in unacceptable heating of the chamber wall or beam instabilities. The impedance measurements involve utilizing both a single displaced wire and two wires excited in opposite phase to disentangle the driving and detuning transverse impedances. Trapped mode resonances and longitudinal impedance are to also be measured and compared with simulations. These measurements, when completed, will demonstrate the device is fully operational and has the impedance characteristics and BPM performance acceptable for installation in the SPS.

  4. Can Collimated Extraterrestrial Signals be Intercepted?

    NASA Astrophysics Data System (ADS)

    Forgan, D. H.

    2014-06-01

    The Optical Search for Extraterrestrial Intelligence (OSETI) attempts to detect collimated, narrow-band pulses of electromagnetic radiation. These pulses may either consist of signals intentionally directed at the Earth, or signals between two star systems with a vector that unintentionally intersects the Solar System, allowing Earth to intercept the communication. But should we expect to be able to intercept these unintentional signals? And what constraints can we place upon the frequency of intelligent civilisations if we do? We carry out Monte Carlo Realisation simulations of interstellar communications between civilisations in the Galactic Habitable Zone (GHZ) using collimated beams. We measure the frequency with which beams between two stars are intercepted by a third. The interception rate increases linearly with the fraction of communicating civilisations, and as the cube of the beam opening angle, which is somewhat stronger than theoretical expectations, which we argue is due to the geometry of the GHZ. We find that for an annular GHZ containing 10,000 civilisations, intersections are unlikely unless the beams are relatively uncollimated. These results indicate that optical SETI is more likely to find signals deliberately directed at the Earth than accidentally intercepting collimated communications. Equally, civilisations wishing to establish a network of communicating species may use weakly collimated beams to build up the network through interception, if they are willing to pay a cost penalty that is lower than that meted by fully isotropic beacons. Future SETI searches should consider the possibility that communicating civilisations will attempt to strike a balance between optimising costs and encouraging contact between civilisations, and look for weakly collimated pulses as well as narrow-beam pulses directed deliberately at the Earth.

  5. Comparison of pinhole collimator materials based on sensitivity equivalence

    NASA Astrophysics Data System (ADS)

    Bom, Victor; Goorden, Marlies; Beekman, Freek

    2011-06-01

    Pinhole SPECT often provides an excellent resolution sensitivity trade-off for radionuclide imaging compared to SPECT with parallel holes, particularly when imaging small experimental animals like rodents. High absorption pinhole materials are often chosen because of their low edge penetration and therefore good system resolution. Capturing more photons in the edges however results in decreased system sensitivity if the pinhole diameter remains the same, which may partly undo the beneficial effect on the resolution. In the search for an optimal trade-off we have compared pinhole projection data and reconstructed images of different materials with pinhole aperture diameters adjusted to obtain equal sensitivity. Monte Carlo calculations modeling the transmission, penetration and scattering of gamma radiation in single pinholes of uranium, gold, tungsten and lead were performed for a range of pinhole opening angles, diameters and gamma ray energies. In addition, reconstructed images of a hot rod phantom were determined for a multipinhole SPECT system and for a system that can image the 511 keV annihilation photons of positron emitting tracers with clustered pinholes. Our results indicate that, under the condition of equal sensitivity, tungsten and for SPECT also lead pinholes perform just as well as gold and uranium ones, indicating that a significant cost reduction can be achieved in pinhole collimator manufacturing while the use of rare or impractical materials can be avoided.

  6. Novel Collimated Beam Setup to Study the Kinetics of VUV-Induced Reactions.

    PubMed

    Duca, Clara; Imoberdorf, Gustavo; Mohseni, Madjid

    2014-01-01

    Vacuum UV (VUV) process is an incipient advanced oxidation process, which can be used for water treatment. This process relies on the formation of hydroxyl radicals through the VUV-induced photolysis of water. In particular, the use of ozone-generating mercury vapor lamps, which emit 10% of the radiation at 185 nm and 90% at 254 nm, is showing very promising results for the degradation of micropollutants. The kinetics of VUV process has been studied in batch- and flow-through reactors, but the effect of 254 and 185 nm photons cannot be isolated, mass transfer resistances can take place and the interpretation of the results is complex. In this technical note, a new VUV collimated beam to conduct kinetic tests is presented, which offers several advantages: (1) it allows the irradiation of samples with 185, 254 nm photons, or both, (2) the concentration of reagents is uniform in the reaction volume and (3) it allows to change the fluence rate by changing the distance between the lamp and the photoreactor. Details of the geometry are presented, as well as an analysis of the collimation and uniformity of the radiation of the new VUV-collimated beam setup. PMID:23952050

  7. PERFORMANCE OF AND UPGRADES TO THE SNS COLLIMATOR SYSTEMS

    SciTech Connect

    Plum, Michael A; Abdou, Ashraf A; Jacobs, Lorelei L; Janney, Jim G; Geoghegan, Patrick J; McTeer, Stephen Mark; Popova, Irina; Ferguson, Phillip D; Zhukov, Alexander P

    2009-01-01

    As the Spallation Neutron Source (SNS) beam power is increased, the collimator systems are becoming correspondingly more important. The High Energy Beam Transport (HEBT) transverse collimators are now routinely used during neutron production. We are in the process of redesigning the HEBT momentum collimation system due to problems with gas production from radiolysis. The Ring collimators are designed for two-stage operation but to date they are mainly used in one-stage mode. In this paper we will discuss the status, the operational performance, and upgrades to the collimation systems.

  8. Performance of a Lanthanum Bromide Detector and a New Conception Collimator for Radiopharmaceuticals Molecular Imaging in Oncology

    SciTech Connect

    Pani, Roberto; Pellegrini, Rosanna; Bennati, Paolo; Cinti, Maria Nerina; Scafe, Raffaele; De Vincentis, Giuseppe; Navarria, Francesco; Moschini, Giuliano; Rossi, Paolo; Cencelli, Valentino Orsolini; De Notaristefani, Francesco

    2009-03-10

    We have realized and tested a new-design compact gamma camera for high resolution SPET (Single Photon Emission Tomography), and small animals' radio-pharmaceutical molecular imaging. The camera is based on a 'continuous' Lanthanum tri-Bromide crystal, and a new Low Energy (LE) collimator. The crystal is interfaced to a 2x2 array of Hamamatsu-H8500 position sensitive photo-multipliers. The lead collimator features parallel hexagonal 1.0 mm holes, 18 mm length, 0.2 mm septa and 10x10 cm{sup 2} detection area. It was newly designed to fully exploit the high spatial resolution a Lanthanum crystal may provide. To better evaluate its role, we have compared our camera to three other systems with similar crystals and photomultipliers, but employing traditional collimators, either pinhole or parallel. The new camera seems to be complementary to pinhole systems and shows a very attractive trade-off between spatial resolution and detection area.

  9. Performance of a Lanthanum Bromide Detector and a New Conception Collimator for Radiopharmaceuticals Molecular Imaging in Oncology

    NASA Astrophysics Data System (ADS)

    Pani, Roberto; Pellegrini, Rosanna; Bennati, Paolo; Cinti, Maria Nerina; Scafè, Raffaele; De Vincentis, Giuseppe; Navarria, Francesco; Moschini, Giuliano; Cencelli, Valentino Orsolini; De Notaristefani, Francesco; Rossi, Paolo

    2009-03-01

    We have realized and tested a new-design compact gamma camera for high resolution SPET (Single Photon Emission Tomography), and small animals' radio-pharmaceutical molecular imaging. The camera is based on a "continuous" Lanthanum tri-Bromide crystal, and a new Low Energy (LE) collimator. The crystal is interfaced to a 2×2 array of Hamamatsu-H8500 position sensitive photo-multipliers. The lead collimator features parallel hexagonal 1.0 mm holes, 18 mm length, 0.2 mm septa and 10×10 cm2 detection area. It was newly designed to fully exploit the high spatial resolution a Lanthanum crystal may provide. To better evaluate its role, we have compared our camera to three other systems with similar crystals and photomultipliers, but employing traditional collimators, either pinhole or parallel. The new camera seems to be complementary to pinhole systems and shows a very attractive trade-off between spatial resolution and detection area.

  10. X-ray collimation by the parabolic cylinder mirror in SPring-8/BL29XUL.

    PubMed

    Takei, Dai; Kohmura, Yoshiki; Senba, Yasunori; Ohashi, Haruhiko; Tamasaku, Kenji; Ishikawa, Tetsuya

    2016-01-01

    A combination of plane and threefold-shape X-ray mirrors was installed in SPring-8 BL29XUL. The second mirror has parabolic cylinder surfaces that collimate X-rays in the vertical direction. A performance test was conducted, yielding highly collimated 8 keV photon beams with an effective angular divergence of 0.4 µrad, below only 5% of that of the original beams. The double-mirror system preserved 70% of the total incident flux and nearly tripled the flux density at 988 m from the light source. The values of the observations were almost similar to those of our ray-tracing simulation. Based on the results a discussion of future prospects of the mirror system is included.

  11. Imaging the sun in hard X-rays - Spatial and rotating modulation collimators

    NASA Technical Reports Server (NTRS)

    Campbell, Jonathan W.; Davis, John M.; Emslie, A. G.

    1991-01-01

    Several approaches to imaging hard X-rays emitted from solar flares have been proposed or are planned for the nineties including the spatial modulation collimator (SMC) and the rotating modulation collimator (RMC). A survey of current solar flare theoretical literature indicates the desirability of spatial resolutions down to 1 arcsecond, field of views greater than the full solar disk (i.e., 32 arcminutes), and temporal resolutions down to 1 second. Although the sun typically provides relatively high flux levels, the requirement for 1 second temporal resolution raises the question as to the viability of Fourier telescopes subject to the aforementioned constraints. A basic photon counting, Monte Carlo 'end-to-end' model telescope was employed using the Astronomical Image Processing System (AIPS) for image reconstruction. The resulting solar flare hard X-ray images compared against typical observations indicated that both telescopes show promise for the future.

  12. TH-C-BRD-03: Determining the Optimal Collimator Position for Collimated Pencil Beam Scanning Proton Therapy

    SciTech Connect

    Wang, D; Smith, B; Hill, P; Gelover, E; Flynn, R; Hyer, D

    2014-06-15

    Purpose: There has been a growing interest in applying collimation to pencil beam scanning (PBS) proton therapy in order to sharpen the lateral dose falloff out of the target, especially at low energies. Currently, there is not a method to optimally determine the collimation position or margin around the target. A uniform margin would not be ideal due to the fact that an incoming symmetric pencil beam, after being intercepted by a collimator near the target boundary, will become asymmetric and experience a lateral shift away from its original spot location, leaving the target insufficiently covered. We demonstrate a method that optimally determines the collimator position on a per-spot basis, in order to maximize target dose while minimizing normal tissue dose. Methods: A library of collimated pencil beams were obtained through Monte Carlo simulation with a collimator placed at varying distances from the central axis of an incoming symmetrical pencil beam of 118 MeV and 5 mm sigma-in-air. Two-dimensional treatment plans were then created using this library of collimated pencil beams. For each spot position in a treatment plan, the collimator position was optimally determined in such a way that the resultant pencil beam maximized the ratio of in-target dose and out-of-target dose. For comparison, un-collimated treatment plans were also computed. Results: The spot-by-spot optimally determined collimator positions allowed the reduction of normal tissue dose while maintaining the same target coverage as un-collimated PBS. Quantitatively, the mean dose outside of the target was reduced by approximately 40% as compared to the plan without collimation. Conclusion: The proposed method determines the optimal collimator position for each spot in collimated PBS proton therapy. The use of a collimator will improve PBS dose distributions achievable today and will continue to be the subject of future investigations.

  13. Electron lenses for particle collimation in LHC

    SciTech Connect

    Shiltsev, v.; /Fermilab

    2007-12-01

    Electron Lenses built and installed in Tevatron have proven themselves as safe and very reliable instruments which can be effectively used in hadron collider operation for a number of applications, including compensation of beam-beam effects [1], DC beam removal from abort gaps [2], as a diagnostic tool. In this presentation we - following original proposal [3] - consider in more detail a possibility of using electron lenses with hollow electron beam for ion and proton collimation in LHC.

  14. Subwavelength nanobrush target to collimate fast electrons

    NASA Astrophysics Data System (ADS)

    Zhao, Zongqing; Cao, Lihua; Cao, Leifeng; Zhou, Weimin; Wu, Yuchi; Zhu, Bin; Dong, Kegong; Zhang, Baohan; Ding, Yongkun; Gu, Yuqiu

    2011-10-01

    A subwavelength nanobrush target was proposed to collimate fast electrons in laser plasma interaction, which consists of a 5 μm copper underlay covered with a 20 μm thick layer of metallic fibers. The diameter of the individual fibers is about 200 nm and the spacing between them is about 150 nm. The experiment was hold at SILEX-I laser facility (10 J, 31 fs, 300 TW). When a subwavelength nanobrush target interacts with ultraintense laser of 7.9*1018/cm2, highly collimated fast electron beam with divergence angle nearly zero whereas the divergence of the plane target is 40 degree. Two-dimensional particle-in-cell (PIC) simulations show that the fast electrons will be accelerated and guided by strong transient electromagnetic fields created at the wall surfaces of nanobrushs. Both experiment and simulation show that the subwavelength nanobrush target can indeed generate fast electrons more efficiency and collimate them. The scheme should be useful for fast ignition and K α source research in inertial confinement fusion.

  15. Characterization of a SPECT pinhole collimator for optimal detector usage (the lofthole)

    NASA Astrophysics Data System (ADS)

    Deprez, Karel; Pato, Lara R. V.; Vandenberghe, Stefaan; Van Holen, Roel

    2013-02-01

    In single-photon emission computed tomography (SPECT), multi-pinhole collimation is often employed nowadays. Most multi-pinhole collimators avoid overlap (multiplexing) of the projections on the detector. This can be done by using additional shielding or by spacing the pinholes far enough apart. Using additional shielding has the drawback that it increases weight, design complexity and cost. Spacing the pinholes far enough apart results in sub-optimal detector usage, the valuable detector area is not entirely used. This is due to the circular projections of pinholes on the detector; these ellipses can not be tiled with high detector coverage. To overcome this we designed a new pinhole geometry, the lofthole, that has a rectangular projection on the detector. The lofthole has a circular aperture and a rectangular entrance/exit opening. Sensitivity formulae have been derived for pinholes and loftholes. These formulae take the penumbra effect into account; the proposed formulae do not take penetration into account. The derived formulae are valid for geometries where the field-of-view and the sensitivity of the aperture are solely limited by the exit window. A flood map measurement was performed to compare the rectangular projection of a lofthole with the circular projection of a pinhole. Finally, measurements were done to compare the amount of penetration of pinholes with the amount of penetration of a lofthole. A square lofthole collimator has less penetration than a knife-edge pinhole collimator that irradiates the same rectangular detector area with full coverage. A multi-lofthole collimator allows high detector coverage without using additional shielding. An additional advantage is the lower amount of penetration.

  16. A Maximum NEC Criterion for Compton Collimation to Accurately Identify True Coincidences in PET

    PubMed Central

    Chinn, Garry; Levin, Craig S.

    2013-01-01

    In this work, we propose a new method to increase the accuracy of identifying true coincidence events for positron emission tomography (PET). This approach requires 3-D detectors with the ability to position each photon interaction in multi-interaction photon events. When multiple interactions occur in the detector, the incident direction of the photon can be estimated using the Compton scatter kinematics (Compton Collimation). If the difference between the estimated incident direction of the photon relative to a second, coincident photon lies within a certain angular range around colinearity, the line of response between the two photons is identified as a true coincidence and used for image reconstruction. We present an algorithm for choosing the incident photon direction window threshold that maximizes the noise equivalent counts of the PET system. For simulated data, the direction window removed 56%–67% of random coincidences while retaining > 94% of true coincidences from image reconstruction as well as accurately extracted 70% of true coincidences from multiple coincidences. PMID:21317079

  17. Source holder collimator for encapsulating radioactive material and collimating the emanations from the material

    DOEpatents

    Laurer, G.R.

    1974-01-22

    This invention provides a transportable device capable of detecting normal levels of a trace element, such as lead in a doughnutshaped blood sample by x-ray fluorescence with a minimum of sample preparation in a relatively short analyzing time. In one embodiment, the blood is molded into a doughnut-shaped sample around an annular array of low-energy radioactive material that is at the center of the doughnut-shaped sample but encapsulated in a collimator, the latter shielding a detector that is close to the sample and facing the same so that the detector receives secondary emissions from the sample while the collimator collimates ths primary emissions from the radioactive material to direct these emissions toward the sample around 360 deg and away from the detector. (Official Gazette)

  18. Treatment planning systems for external whole brain radiation therapy: With and without MLC (multi leaf collimator) optimization

    NASA Astrophysics Data System (ADS)

    Budiyono, T.; Budi, W. S.; Hidayanto, E.

    2016-03-01

    Radiation therapy for brain malignancy is done by giving a dose of radiation to a whole volume of the brain (WBRT) followed by a booster at the primary tumor with more advanced techniques. Two external radiation fields given from the right and left side. Because the shape of the head, there will be an unavoidable hotspot radiation dose of greater than 107%. This study aims to optimize planning of radiation therapy using field in field multi-leaf collimator technique. A study of 15 WBRT samples with CT slices is done by adding some segments of radiation in each field of radiation and delivering appropriate dose weighting using a TPS precise plan Elekta R 2.15. Results showed that this optimization a more homogeneous radiation on CTV target volume, lower dose in healthy tissue, and reduced hotspots in CTV target volume. Comparison results of field in field multi segmented MLC technique with standard conventional technique for WBRT are: higher average minimum dose (77.25% ± 0:47%) vs (60% ± 3:35%); lower average maximum dose (110.27% ± 0.26%) vs (114.53% ± 1.56%); lower hotspot volume (5.71% vs 27.43%); and lower dose on eye lenses (right eye: 9.52% vs 18.20%); (left eye: 8.60% vs 16.53%).

  19. Hybrid algorithm for simulating the collimated transmittance of homogeneous stratified turbid media

    PubMed Central

    Cruzado, Beatriz Morales; Atencio, José Alberto Delgado; Vázquez y Montiel, Sergio; Gómez, Erick Sarmiento

    2015-01-01

    In this work we describe the development of a program that simulates the propagation of photons through refractive and reflecting optical components such as lenses, mirrors and stops that includes a biological tissue sample as the main issue to be investigated in order to get a simulated value of light distribution, in particular, of the unscattered light. The analysis of the photons that travel through the sample is based on the program Monte Carlo Multi-Layered with some modifications that consider a Gaussian beam as initial source of light. Position, directional cosines and weight of photons exiting the turbid media are used to propagate them through an optical system. As a mean of validation of the program, we selected a typical optical system for measurement of collimated transmittance. Therefore, several tests were carried out to find the optical system that gives the theoretical collimated transmittance at different values of the optical properties of the turbid media. Along this validation, the optimal experimental configuration is found. Using this results, a comparison between the simulated optimal configuration and the experimental set-up was done, by using a colloidal suspension as a turbid media. PMID:26137375

  20. The influence of penetrating radiation on collimator performance.

    PubMed

    Jahns, M J

    1981-01-01

    Geometric and penetration responses of a multihole focusing collimator to a point source have been determined using a digital computer and a ray tracing technique. Results are presented as point and line spread functions for collimated rays and penetrating rays, and as plane sensitivities and modulation transfer functions derived from the line spread functions. Responses to a 141Ce line source have been measured for a lead collimator and for a cadium collimator of practically identical dimensions. A comparison of measured and computed responses shows that while computed penetration approximates measured penetration reasonably well, there is a measurable contribution from scattered rays. Plane sensitivities and the modulation transfer function derived from the computed responses are shown. The effects of penetration are compared with predictions made by other investigators. This analysis of collimator performance provides information not previously available and should be useful for evaluating collimators of existing or proposed design.

  1. Collimator wake fields in the SLC final focus

    SciTech Connect

    Zimmermann, F.; Bane, K.L.F.; Ng, C.K.

    1996-06-01

    The SLC final focus system accommodates 29 fixed or adjustable collimators for machine protection and background reduction. By amplifying pulse to pulse orbit variations and by generating emittance growth, collimator wake fields may degrade the beam quality at the interaction point (IP). In the SLC final focus, collimator half apertures are larger than the bunch length, so that the standard collimator wake formula of Bane and Morton does not apply. Numerical wake field calculations for SLC parameters agree quite well with the high frequency impedance of a step out transition. Due to the nature of a final focus system, the wake field kicks from all collimators add coherently, and the overall impact on luminosity can be significant. This paper suggests that collimator wake fields in the final focus provide a possible explanation for the 30% discrepancy between expected and measured luminosity in the 1994/95 SLC run.

  2. Collimator application for microchannel plate image intensifier resolution improvement

    DOEpatents

    Thomas, Stanley W.

    1996-02-27

    A collimator is included in a microchannel plate image intensifier (MCPI). Collimators can be useful in improving resolution of MCPIs by eliminating the scattered electron problem and by limiting the transverse energy of electrons reaching the screen. Due to its optical absorption, a collimator will also increase the extinction ratio of an intensifier by approximately an order of magnitude. Additionally, the smooth surface of the collimator will permit a higher focusing field to be employed in the MCP-to-collimator region than is currently permitted in the MCP-to-screen region by the relatively rough and fragile aluminum layer covering the screen. Coating the MCP and collimator surfaces with aluminum oxide appears to permit additional significant increases in the field strength, resulting in better resolution.

  3. Collimator application for microchannel plate image intensifier resolution improvement

    DOEpatents

    Thomas, S.W.

    1996-02-27

    A collimator is included in a microchannel plate image intensifier (MCPI). Collimators can be useful in improving resolution of MCPIs by eliminating the scattered electron problem and by limiting the transverse energy of electrons reaching the screen. Due to its optical absorption, a collimator will also increase the extinction ratio of an intensifier by approximately an order of magnitude. Additionally, the smooth surface of the collimator will permit a higher focusing field to be employed in the MCP-to-collimator region than is currently permitted in the MCP-to-screen region by the relatively rough and fragile aluminum layer covering the screen. Coating the MCP and collimator surfaces with aluminum oxide appears to permit additional significant increases in the field strength, resulting in better resolution. 2 figs.

  4. Beam Delivery WG Summary: Optics, Collimation & Background

    SciTech Connect

    Angal-Kalinin, D.; Jackson, F.; Mokhov, N.V.; Kuroda, S.; Seryi, A.A.; /SLAC

    2006-01-20

    The presented paper partially summarizes the work of the Beam Delivery working group (WG4) at Snowmass, concentrating on status of optics, layout, collimation, and background. The strawman layout with 2 interaction regions was recommended at the first ILC workshop at KEK in November 2004. Two crossing-angle designs were included in this layout. The design of the ILC BDS has evolved since the first ILC workshop. The progress on the BDS design and extraction line design has been reviewed and the design issues were discussed during the optics and layout session at the Snowmass.

  5. LHC particle collimation with hollow electron beams

    SciTech Connect

    Shiltsev, V.; Drozhdin, A.; Kamerdzhiev, V.; Kuznetsov, G.; Vorobiev, L.; /Fermilab

    2008-06-01

    Electron lenses built and installed in the Tevatron have proven themselves as safe and very reliable instruments which can be effectively used in hadron collider operation for a number of applications, including compensation of beam-beam effects [1], a DC beam removal from abort gaps [2], and as a versatile diagnostic tool. In this article, we--following the original proposal [3,4]--consider in more detail a possibility of using electron lenses with hollow electron beam for ion and proton collimation in LHC and the Tevatron.

  6. Ultra-compact on-chip LED collimation optics by 3D femtosecond direct laser writing.

    PubMed

    Thiele, Simon; Gissibl, Timo; Giessen, Harald; Herkommer, Alois M

    2016-07-01

    By using two-photon lithographic 3D printing, we demonstrate additive manufacturing of a dielectric concentrator directly on a LED chip. With a size of below 200 μm in diameter and length, light output is increased by a factor of 6.2 in collimation direction, while the emission half-angle is reduced by 50%. We measure excellent form fidelity and irradiance patterns close to simulation. Additionally, a more complex shape design is presented, which exhibits a nonconventional triangular illumination pattern. The introduced method features exceptional design freedoms which can be used to tailor high-quality miniature illumination optics for specific lighting tasks, for example, endoscopy. PMID:27367093

  7. Ballistic imaging of biological media with collimated illumination and focal plane detection

    NASA Astrophysics Data System (ADS)

    Brezner, Barak; Cahen, Sarah; Glasser, Ziv; Sternklar, Shmuel; Granot, Er'el

    2015-07-01

    A simple, affordable method for imaging through biological tissue is investigated. The method consists of (1) imaging with a wavelength that has a relatively small scattering coefficient (1310 nm in this case) and (2) collimated illumination together with (3) focal plane detection to enhance the detection of the ballistic photons relative to the diffusive light. We demonstrate ballistic detection of an object immersed in a 1-cm-thick cuvette filled with 4% Intralipid, which is equivalent to ˜1 to 2 cm of skin tissue. With the same technology, a ballistic image of a 1-mm-wide object in 10-mm-thick chicken breast is also presented.

  8. Collimator and energy window optimization for ⁹⁰Y bremsstrahlung SPECT imaging: A SIMIND Monte Carlo study.

    PubMed

    Roshan, Hoda Rezaei; Mahmoudian, Babak; Gharepapagh, Esmaeil; Azarm, Ahmadreza; Islamian, Jalil Pirayesh

    2016-02-01

    Treatment efficacy of radioembolization using Yttrium-90 ((90)Y) microspheres is assessed by the (90)Y bremsstrahlung single photon emission computed tomography (SPECT) imaging following radioembolization. The radioisotopic image has the potential of providing reliable activity map of (90)Y microspheres distribution. One of the main reasons of the poor image quality in (90)Y bremsstrahlung SPECT imaging is the continuous and broad energy spectrum of the related bremsstrahlung photons. Furthermore, collimator geometry plays an impressive role in the spatial resolution, sensitivity and image contrast. Due to the relatively poor quality of the (90)Y bremsstrahlung SPECT images, we intend to optimize the medium-energy (ME) parallel-hole collimator and energy window. The Siemens e.cam gamma camera equipped with a ME collimator and a voxelized phantom was simulated by the SImulating Medical Imaging Nuclear Detectors (SIMIND) program. We used the SIMIND Monte Carlo program to generate the (90)Y bremsstrahlung SPECT projection of the digital Jaszczak phantom. The phantom consist of the six hot spheres ranging from 9.5 to 31.8mm in diameter, which are used to evaluate the image contrast. In order to assess the effect of the energy window on the image contrast, three energy windows ranging from 60 to 160 KeV, 160 to 400 KeV, and 60 to 400 KeV were set on a (90)Y bremsstrahlung spectrum. As well, the effect of the hole diameter of a ME collimator on the image contrast and bremsstrahlung spectrum were investigated. For the fixed collimator and septa thickness values (3.28 cm and 1.14 mm, respectively), a hole diameter range (2.35-3.3mm) was chosen based on the appropriate balance between the spatial resolution and sensitivity. The optimal energy window for (90)Y bremsstrahlung SPECT imaging was extended energy window from 60 to 400 KeV. Besides, The optimal value of the hole diameter of ME collimator was obtained 3.3mm. Geometry of the ME parallel-hole collimator and energy

  9. Design of optimal collimation for dedicated molecular breast imaging systems

    SciTech Connect

    Weinmann, Amanda L.; Hruska, Carrie B.; O'Connor, Michael K.

    2009-03-15

    Molecular breast imaging (MBI) is a functional imaging technique that uses specialized small field-of-view gamma cameras to detect the preferential uptake of a radiotracer in breast lesions. MBI has potential to be a useful adjunct method to screening mammography for the detection of occult breast cancer. However, a current limitation of MBI is the high radiation dose (a factor of 7-10 times that of screening mammography) associated with current technology. The purpose of this study was to optimize the gamma camera collimation with the aim of improving sensitivity while retaining adequate resolution for the detection of sub-10-mm lesions. Square-hole collimators with holes matched to the pixilated cadmium zinc telluride detector elements of the MBI system were designed. Data from MBI patient studies and parameters of existing dual-head MBI systems were used to guide the range of desired collimator resolutions, source-to-collimator distances, pixel sizes, and collimator materials that were examined. General equations describing collimator performance for a conventional gamma camera were used in the design process along with several important adjustments to account for the specialized imaging geometry of the MBI system. Both theoretical calculations and a Monte Carlo model were used to measure the geometric efficiency (or sensitivity) and resolution of each designed collimator. Results showed that through optimal collimation, collimator sensitivity could be improved by factors of 1.5-3.2, while maintaining a collimator resolution of either {<=}5 or {<=}7.5 mm at a distance of 3 cm from the collimator face. These gains in collimator sensitivity permit an inversely proportional drop in the required dose to perform MBI.

  10. Mechanical approach to the neutrons spectra collimation and detection

    SciTech Connect

    Sadeghi, H.; Roshan, M. V.

    2014-11-15

    Neutrons spectra from most of known sources require being collimated for numerous applications; among them one is the Neutron Activation Analysis. High energy neutrons are collimated through a mechanical procedure as one of the most promising methods. The output energy of the neutron beam depends on the velocity of the rotating Polyethylene disks. The collimated neutrons are then measured by an innovative detection technique with high accuracy.

  11. Some basic considerations of measurements involving collimated direct sunlight

    NASA Technical Reports Server (NTRS)

    Chai, A. T.

    1976-01-01

    The geometry of collimators for devices or instruments dealing with terrestrial direct sunlight is discussed. Effects of the opening angle and slope angle of a collimator on the measurements are investigated with regard to variations of turbidity and air mass. Based on this investigation, geometric dimensions for collimators and certain realistic terrestrial reference conditions are recommended for the purpose of solar cell calibration in terrestrial applications.

  12. Mechanical approach to the neutrons spectra collimation and detection.

    PubMed

    Sadeghi, H; Roshan, M V

    2014-11-01

    Neutrons spectra from most of known sources require being collimated for numerous applications; among them one is the Neutron Activation Analysis. High energy neutrons are collimated through a mechanical procedure as one of the most promising methods. The output energy of the neutron beam depends on the velocity of the rotating Polyethylene disks. The collimated neutrons are then measured by an innovative detection technique with high accuracy.

  13. Photon-photon collisions

    SciTech Connect

    Burke, D.L.

    1982-10-01

    Studies of photon-photon collisions are reviewed with particular emphasis on new results reported to this conference. These include results on light meson spectroscopy and deep inelastic e..gamma.. scattering. Considerable work has now been accumulated on resonance production by ..gamma gamma.. collisions. Preliminary high statistics studies of the photon structure function F/sub 2//sup ..gamma../(x,Q/sup 2/) are given and comments are made on the problems that remain to be solved.

  14. Photon-photon colliders

    SciTech Connect

    Sessler, A.M.

    1995-04-01

    Since the seminal work by Ginsburg, et at., the subject of giving the Next Linear Collider photon-photon capability, as well as electron-positron capability, has drawn much attention. A 1990 article by V.I. Teinov describes the situation at that time. In March 1994, the first workshop on this subject was held. This report briefly reviews the physics that can be achieved through the photon-photon channel and then focuses on the means of achieving such a collider. Also reviewed is the spectrum of backscattered Compton photons -- the best way of obtaining photons. We emphasize the spectrum actually obtained in a collider with both polarized electrons and photons (peaked at high energy and very different from a Compton spectrum). Luminosity is estimated for the presently considered colliders, and interaction and conversion-point geometries are described. Also specified are laser requirements (such as wavelength, peak power, and average power) and the lasers that might be employed. These include conventional and free-electron lasers. Finally, we describe the R&D necessary to make either of these approaches viable and explore the use of the SLC as a test bed for a photon-photon collider of very high energy.

  15. Dosimetric characterization of a bi-directional micromultileaf collimator for stereotactic applications.

    PubMed

    Bucciolini, M; Russo, S; Banci Buonamici, F; Pini, S; Silli, P

    2002-07-01

    A 6 MV photon beam from Linac SL75-5 has been collimated with a new micromultileaf device that is able to shape the field in the two orthogonal directions with four banks of leaves. This is the first clinical installation of the collimator and in this paper the dosimetric characterization of the system is reported. The dosimetric parameters required by the treatment planning system used for the dose calculation in the patient are: tissue maximum ratios, output factors, transmission and leakage of the leaves, penumbra values. Ionization chambers, silicon diode, radiographic films, and LiF thermoluminescent dosimeters have been employed for measurements of absolute dose and beam dosimetric data. Measurements with different dosimeters supply results in reasonable agreement among them and consistent with data available in literature for other models of micromultileaf collimator; that permits the use of the measured parameters for clinical applications. The discrepancies between results obtained with the different detectors (around 2%) for the analyzed parameters can be considered an indication of the accuracy that can be reached by current stereotactic dosimetry.

  16. Leaky wave lenses for spoof plasmon collimation.

    PubMed

    Panaretos, Anastasios H; Werner, Douglas H

    2016-06-27

    We theoretically demonstrate the feasibility of collimating radiating spoof plasmons using a leaky wave lens approach. Spoof plasmons are surface waves excited along reactance surfaces realized through metallic corrugations. By employing a periodic perturbation to the geometric profile of this type of reactance surface, it becomes feasible to convert the excited spoof plasmons into free-space radiating leaky wave modes. It is demonstrated that by structurally modifying such a corrugated surface through the introduction of a non-uniform sinusoidally modulated reactance profile, then a tapered wavenumber, with a real part less than that of free space, can be established along the surface. In this way the radiating properties of the structure (amplitude and phase) can be locally controlled thereby creating a radiating effect similar to that of a non-uniform current distribution. By properly engineering the space dependent wavenumber along the corrugated surface, different regions of the structure will emit spoof plasmon energy at different angles with varying intensity. The combined effect is the emission of an electromagnetic wave exhibiting a converging wave-front that eventually collimates spoof plasmon energy at some desired focal point.

  17. A variable-collimation display system

    NASA Astrophysics Data System (ADS)

    Batchko, Robert; Robinson, Sam; Schmidt, Jack; Graniela, Benito

    2014-03-01

    Two important human depth cues are accommodation and vergence. Normally, the eyes accommodate and converge or diverge in tandem; changes in viewing distance cause the eyes to simultaneously adjust both focus and orientation. However, ambiguity between accommodation and vergence cues is a well-known limitation in many stereoscopic display technologies. This limitation also arises in state-of-the-art full-flight simulator displays. In current full-flight simulators, the out-the-window (OTW) display (i.e., the front cockpit window display) employs a fixed collimated display technology which allows the pilot and copilot to perceive the OTW training scene without angular errors or distortions; however, accommodation and vergence cues are limited to fixed ranges (e.g., ~ 20 m). While this approach works well for long-range, the ambiguity of depth cues at shorter range hinders the pilot's ability to gauge distances in critical maneuvers such as vertical take-off and landing (VTOL). This is the first in a series of papers on a novel, variable-collimation display (VCD) technology that is being developed under NAVY SBIR Topic N121-041 funding. The proposed VCD will integrate with rotary-wing and vertical take-off and landing simulators and provide accurate accommodation and vergence cues for distances ranging from approximately 3 m outside the chin window to ~ 20 m. A display that offers dynamic accommodation and vergence could improve pilot safety and training, and impact other applications presently limited by lack of these depth cues.

  18. Fast 3D iterative image reconstruction for SPECT with rotating slat collimators

    NASA Astrophysics Data System (ADS)

    Van Holen, Roel; Vandenberghe, Stefaan; Staelens, Steven; DeBeenhouwer, Jan; Lemahieu, Ignace

    2009-02-01

    As an alternative to the use of traditional parallel hole collimators, SPECT imaging can be performed using rotating slat collimators. While maintaining the spatial resolution, a gain in image quality could be expected from the higher photon collection efficiency of this type of collimator. However, the use of iterative methods to do fully three-dimensional (3D) reconstruction is computationally much more expensive and furthermore involves slow convergence compared to a classical SPECT reconstruction. It has been proposed to do 3D reconstruction by splitting the system matrix into two separate matrices, forcing the reconstruction to first estimate the sinograms from the rotating slat SPECT data before estimating the image. While alleviating the computational load by one order of magnitude, this split matrix approach would result in fast computation of the projections in an iterative algorithm, but does not solve the problem of slow convergence. There is thus a need for an algorithm which speeds up convergence while maintaining image quality for rotating slat collimated SPECT cameras. Therefore, we developed a reconstruction algorithm based on the split matrix approach which allows both a fast calculation of the forward and backward projection and a fast convergence. In this work, an algorithm of the maximum likelihood expectation maximization (MLEM) type, obtained from a split system matrix MLEM reconstruction, is proposed as a reconstruction method for rotating slat collimated SPECT data. Here, we compare this new algorithm to the conventional split system matrix MLEM method and to a gold standard fully 3D MLEM reconstruction algorithm on the basis of computational load, convergence and contrast-to-noise. Furthermore, ordered subsets expectation maximization (OSEM) implementations of these three algorithms are compared. Calculation of computational load and convergence for the different algorithms shows a speedup for the new method of 38 and 426 compared to the

  19. Photon-photon colliders

    SciTech Connect

    Sessler, Andrew M.

    1996-01-01

    Since the seminal work by Ginsburg, et al., the subject of giving the Next Linear Collider photon-photon capability, as well as electron-positron capability, has drawn much attention [1]. A 1990 article by V.I. Telnov describes the situation at that time [2]. In March 1994, the first workshop on this subject was held [3]. This report briefly reviews the physics that can be achieved through the photon-photon channel and then focuses on the means of achieving such a collider. Also reviewed is the spectrum of backscattered Compton photons—the best way of obtaining photons. We emphasize the spectrum actually obtained in a collider with both polarized electrons and photons (peaked at high energy and very different from a Compton spectrum). Luminosity is estimated for the presently considered colliders, and interaction and conversion-point geometries are described. Also specified are laser requirements (such as wavelength, peak power, and average power) and the lasers that might be employed. These include conventional and free-electron lasers. Finally, we describe the R&D necessary to make either of these approaches viable and explore the use of the SLC as a test bed for a photon-photon collider of very high energy.

  20. Photon-photon collisions

    SciTech Connect

    Brodsky, S.J.

    1988-07-01

    Highlights of the VIIIth International Workshop on Photon-Photon Collisions are reviewed. New experimental and theoretical results were reported in virtually every area of ..gamma gamma.. physics, particularly in exotic resonance production and tests of quantum chromodynamics where asymptotic freedom and factorization theorems provide predictions for both inclusive and exclusive ..gamma gamma.. reactions at high momentum transfer. 73 refs., 12 figs.

  1. Comparison of secondary neutron dose in proton therapy resulting from the use of a tungsten alloy MLC or a brass collimator system

    SciTech Connect

    Diffenderfer, Eric S.; Ainsley, Christopher G.; Kirk, Maura L.; McDonough, James E.; Maughan, Richard L.

    2011-11-15

    Purpose: To apply the dual ionization chamber method for mixed radiation fields to an accurate comparison of the secondary neutron dose arising from the use of a tungsten alloy multileaf collimator (MLC) as opposed to a brass collimator system for defining the shape of a therapeutic proton field. Methods: Hydrogenous and nonhydrogenous ionization chambers were constructed with large volumes to enable measurements of absorbed doses below 10{sup -4} Gy in mixed radiation fields using the dual ionization chamber method for mixed-field dosimetry. Neutron dose measurements were made with a nominal 230 MeV proton beam incident on a closed tungsten alloy MLC and a solid brass block. The chambers were cross-calibrated against a {sup 60}Co-calibrated Farmer chamber in water using a 6 MV x-ray beam and Monte Carlo simulations were performed to account for variations in ionization chamber response due to differences in secondary neutron energy spectra. Results: The neutron and combined proton plus {gamma}-ray absorbed doses are shown to be nearly equivalent downstream from either a closed tungsten alloy MLC or a solid brass block. At 10 cm downstream from the distal edge of the collimating material the neutron dose from the closed MLC was (5.3 {+-} 0.4) x 10{sup -5} Gy/Gy. The neutron dose with brass was (6.4 {+-} 0.7) x 10{sup -5} Gy/Gy. Further from the secondary neutron source, at 50 cm, the neutron doses remain close for both the MLC and brass block at (6.9 {+-} 0.6) x 10{sup -6} Gy/Gy and (6.3 {+-} 0.7) x 10{sup -6} Gy/Gy, respectively. Conclusions: The dual ionization chamber method is suitable for measuring secondary neutron doses resulting from proton irradiation. The results of measurements downstream from a closed tungsten alloy MLC and a brass block indicate that, even in an overly pessimistic worst-case scenario, secondary neutron production in a tungsten alloy MLC leads to absorbed doses that are nearly equivalent to those seen from brass collimators. Therefore

  2. Dispersion in photonic crystals

    NASA Astrophysics Data System (ADS)

    Witzens, Jeremy

    2005-11-01

    Investigations on the dispersive properties of photonic crystals, modified scattering in ring-resonators, monolithic integration of vertical-cavity surface-emitting lasers and advanced data processing techniques for the finite-difference time-domain method are presented. Photonic crystals are periodic mesoscopic arrays of scatterers that modify the propagation properties of electromagnetic waves in a similar way as "natural" crystals modify the properties of electrons in solid-state physics. In this thesis photonic crystals are implemented as planar photonic crystals, i.e., optically thin semiconductor films with periodic arrays of holes etched into them, with a hole-to-hole spacing of the order of the wavelength of light in the dielectric media. Photonic crystals can feature forbidden frequency ranges (the band-gaps) in which light cannot propagate. Even though most work on photonic crystals has focused on these band-gaps for application such as confinement and guiding of light, this thesis focuses on the allowed frequency regions (the photonic bands) and investigates how the propagation of light is modified by the crystal lattice. In particular the guiding of light in bulk photonic crystals in the absence of lattice defects (the self-collimation effect) and the angular steering of light in photonic crystals (the superprism effect) are investigated. The latter is used to design a planar lightwave circuit for frequency domain demultiplexion. Difficulties such as efficient insertion of light into the crystal are resolved and previously predicted limitations on the resolution are circumvented. The demultiplexer is also fabricated and characterized. Monolithic integration of vertical-cavity surface-emitting lasers by means of resonantly enhanced grating couplers is investigated. The grating coupler is designed to bend light through a ninety-degree angle and is characterized with the finite-difference time-domain method. The vertical-cavity surface-emitting lasers are

  3. Collimator design for a multipinhole brain SPECT insert for MRI

    SciTech Connect

    Van Audenhaege, Karen; Van Holen, Roel; Vanhove, Christian; Vandenberghe, Stefaan

    2015-11-15

    Purpose: Brain single photon emission computed tomography (SPECT) imaging is an important clinical tool, with unique tracers for studying neurological diseases. Nowadays, most commercial SPECT systems are combined with x-ray computed tomography (CT) in so-called SPECT/CT systems to obtain an anatomical background for the functional information. However, while CT images have a high spatial resolution, they have a low soft-tissue contrast, which is an important disadvantage for brain imaging. Magnetic resonance imaging (MRI), on the other hand, has a very high soft-tissue contrast and does not involve extra ionizing radiation. Therefore, the authors designed a brain SPECT insert that can operate inside a clinical MRI. Methods: The authors designed and simulated a compact stationary multipinhole SPECT insert based on digital silicon photomultiplier detector modules, which have shown to be MR-compatible and have an excellent intrinsic resolution (0.5 mm) when combined with a monolithic 2 mm thick LYSO crystal. First, the authors optimized the different parameters of the SPECT system to maximize sensitivity for a given target resolution of 7.2 mm in the center of the field-of-view, given the spatial constraints of the MR system. Second, the authors performed noiseless simulations of two multipinhole configurations to evaluate sampling and reconstructed resolution. Finally, the authors performed Monte Carlo simulations and compared the SPECT insert with a clinical system with ultrahigh-resolution (UHR) fan beam collimators, based on contrast-to-noise ratio and a visual comparison of a Hoffman phantom with a 9 mm cold lesion. Results: The optimization resulted in a stationary multipinhole system with a collimator radius of 150.2 mm and a detector radius of 172.67 mm, which corresponds to four rings of 34 diSPM detector modules. This allows the authors to include eight rings of 24 pinholes, which results in a system volume sensitivity of 395 cps/MBq. Noiseless simulations

  4. Evaluation of collimation and imaging configuration in scintimammography

    SciTech Connect

    Tsui, B.M.W.; Frey, E.C.; Wessell, D.E.

    1996-12-31

    Conventional scintimammography (SM) with {sup 99m}Tc sestamibi has been limited to taking a single lateral view of the breast using a parallel-hole high resolution (LEHR) collimator. The collimator is placed close to the breast for best possible spatial resolution. However, the collimator geometry precludes imaging the breast from other views. We evaluated using a pinhole collimator instead of a LEHR collimator in SM for improved spatial resolution and detection efficiency, and to allow additional imaging views. Results from theoretical calculations indicated that pinhole collimators could be designed with higher spatial resolution and detection efficiency than LEHR when imaging small to medium size breasts. The geometrical shape of the pinhole collimator allows imaging of the breasts from both the lateral and craniocaudal views. The dual-view images allow better determination of the location of the tumors within the breast and improved detection of tumors located in the medial region of the breast. A breast model that simulates the shape and composition of the breast and breast tumors with different sizes and locations was added to an existing 3D mathematical cardiac-torso (MCAT) phantom. A cylindrically shaped phantom with 10 cm diameter and spherical inserts with different sizes and {sup 99m}Tc sestamibi uptakes with respect to the background provide physical models of breast with tumors. Simulation studies using the breast and MCAT phantoms and experimental studies using the cylindrical phantom confirmed the utility of the pinhole collimator in SM for improved breast tumor detection.

  5. Self-Collimating Unstable-Resonator Diode Lasers

    NASA Technical Reports Server (NTRS)

    Lang, Robert J.

    1992-01-01

    Proposed unstable-resonator diode lasers produce solid, collimated output beams without external collimating optics. Simpler to construct, used as pumps for optical-fiber amplifiers in telecommunications, for other solid-state lasers, and sources of light for free-space communications.

  6. Self-collimated unstable resonator semiconductor laser

    NASA Technical Reports Server (NTRS)

    Lang, Robert J. (Inventor)

    1993-01-01

    Self-collimation of the output is achieved in an unstable resonator semiconductor laser by providing a large concave mirror M sub 1 and a small convex mirror M sub 2 on opposite surfaces of a semiconductor body of a material having an effective index of refraction denoted by n, where the respective mirror radii R sub 1, R sub 2 and beam radii r sub 1, r sub 2 are chosen to satisfy a condition (R sub 2)/(1 + r sub 1) = (n - 1)/n, with a value of geometric magnification 1 less than or equal to M less than or equal to (n + 1)/(n - 1) where r sub 1 and r sub 2 are the radii of counterpropagating beams at respective mirrors of radii R sub 1 and R sub 2.

  7. Self-collimated unstable resonator semiconductor laser

    NASA Astrophysics Data System (ADS)

    Lang, Robert J.

    1991-08-01

    Self-collimation of the output is achieved in an unstable resonator semiconductor laser by providing a large concave mirror (M sub 1) and a small convex mirror (M sub 2) on opposite surfaces of a semiconductor body of a material having an effective index of refraction denoted by n, where the respective mirror radii (R sub 1, R sub 2) and beam radii (r sub 1, r sub 2) are chosen to satisfy a condition (R sub 2)/(1+r sub 1) = (n-1)/n, with a value of geometric magnification 1 less than or equal to M less than or equal to (n+1)/(n-1) where r (sub 1) and r (sub 2) are the radii of counterpropagating beams at respective mirrors of radii R (sub 1) and R (sub 2).

  8. Beam diffusion measurements using collimator scans in the LHC

    NASA Astrophysics Data System (ADS)

    Valentino, Gianluca; Aßmann, Ralph; Bruce, Roderik; Burkart, Florian; Previtali, Valentina; Redaelli, Stefano; Salvachua, Belen; Stancari, Giulio; Valishev, Alexander

    2013-02-01

    The time evolution of beam losses during a collimator scan provides information on halo diffusion and population. This is an essential input for machine performance characterization and for the design of collimation systems. Beam halo measurements in the CERN Large Hadron Collider were conducted through collimator scrapings in a dedicated beam study for the first time at 4 TeV. Four scans were performed with two collimators, in the vertical plane for beam 1 and horizontally for beam 2, before and after bringing the beams into collisions. Inward and outward steps were performed. A diffusion model was used to interpret the observed loss rate evolution in response to the collimator steps. With this technique, diffusion coefficients were estimated as a function of betatron oscillation amplitude from approximately 3 to 7 standard deviations of the transverse beam distribution. A comparison of halo diffusion and core emittance growth rates is also presented.

  9. Image Reconstruction for a Partially Collimated Whole Body PET Scanner.

    PubMed

    Alessio, Adam M; Schmitz, Ruth E; Macdonald, Lawrence R; Wollenweber, Scott D; Stearns, Charles W; Ross, Steven G; Ganin, Alex; Lewellen, Thomas K; Kinahan, Paul E

    2008-06-01

    Partially collimated PET systems have less collimation than conventional 2-D systems and have been shown to offer count rate improvements over 2-D and 3-D systems. Despite this potential, previous efforts have not established image-based improvements with partial collimation and have not customized the reconstruction method for partially collimated data. This work presents an image reconstruction method tailored for partially collimated data. Simulated and measured sensitivity patterns are presented and provide a basis for modification of a fully 3-D reconstruction technique. The proposed method uses a measured normalization correction term to account for the unique sensitivity to true events. This work also proposes a modified scatter correction based on simulated data. Measured image quality data supports the use of the normalization correction term for true events, and suggests that the modified scatter correction is unnecessary.

  10. [Influence of collimation system on static Fourier transform spectrometer].

    PubMed

    Jiang, Cheng-Zhi; Liang, Jing-Qiu; Liang, Zhong-Zhu; Sun, Qiang; Wang, Wei-Biao

    2014-01-01

    Collimation system provides collimated light for the static Fourier-transform spectroscopy (SFTS). Its quality is crucial to the signal to noise ratio (SNR) of SFTS. In the present paper, the physical model of SFTS was established based on the Fresnel diffraction theory by means of numerical software. The influence of collimation system on the SFTS was discussed in detail focusing on the aberrations of collimation lens and the quality of extended source. The results of simulation show that the influences of different kinds of aberrations on SNR take on obvious regularity, and in particular, the influences of off-axis aberrations on SNR are closely related to the location of off-axis point source. Finally the extended source's maximum radius allowed was obtained by simulation, which equals to 0.65 mm. The discussion results will be used for the design of collimation system.

  11. Preliminary Exploratory Study of Different Phase II Collimators

    SciTech Connect

    Lari, L.; Assmann, R.W.; Bertarelli, A.; Bracco, C.; Brugger, M.; Cerutti, F.; Dallocchio, A.; Ferrari, A.; Mauri, M.; Roesler, S.; Sarchiapone, L.; Vlachoudis, Vasilis; Doyle, J.E.; Keller, L.; Lundgren, S.A.; Markiewicz, Thomas W.; Smith, J.C.; Lari, L.; /LPHE, Lausanne

    2011-11-02

    The Large Hadron Collider (LHC) collimation system is installed and commissioned in different phases, following the natural evolution of the LHC performance. To improve cleaning efficiency towards the end of the low beta squeeze at 7TeV, and in stable physics conditions, it is foreseen to complement the 30 highly robust Phase I secondary collimators with low impedance Phase II collimators. At this stage, their design is not yet finalized. Possible options include metallic collimators, graphite jaws with a movable metallic foil, or collimators with metallic rotating jaws. As part of the evaluation of the different designs, the FLUKA Monte Carlo code is extensively used for calculating energy deposition and studying material damage and activation. This report outlines the simulation approach and defines the critical quantities involved.

  12. MEASUREMENT OF RESISTIVITY DOMINATED COLLIMATOR WAKEFIELD KICKS AT THE SLC

    SciTech Connect

    Onoprienko, D.

    2003-10-03

    Collimation of the multi-MW beams in present linear collider designs is a critical issue. Graphite as a possible collimator material exhibits good mechanical properties but a rather poor electrical conductivity. The resulting resistive wakefields can potentially deteriorate the beam quality. The knowledge of strength and scalings of such wakefields is essential for the design of the collimation systems in linear colliders. We present measurements of the self induced wakefield kick that a charged particle beam experiences when it passes through a narrow gap of two graphite collimator jaws. The measurements where performed with the highly damped SLC beam at an energy of 1.19 GeV. The results are compared to theoretical predictions and to measurements with copper collimators of the same shape.

  13. Estimation of photoneutron yield in linear accelerator with different collimation systems by Geant4 and MCNPX simulation codes

    NASA Astrophysics Data System (ADS)

    Kim, Yoon Sang; Khazaei, Zeinab; Ko, Junho; Afarideh, Hossein; Ghergherehchi, Mitra

    2016-04-01

    At present, the bremsstrahlung photon beams produced by linear accelerators are the most commonly employed method of radiotherapy for tumor treatments. A photoneutron source based on three different energies (6, 10 and 15 MeV) of a linac electron beam was designed by means of Geant4 and Monte Carlo N-Particle eXtended (MCNPX) simulation codes. To obtain maximum neutron yield, two arrangements for the photo neutron convertor were studied: (a) without a collimator, and (b) placement of the convertor after the collimator. The maximum photon intensities in tungsten were 0.73, 1.24 and 2.07 photon/e at 6, 10 and 15 MeV, respectively. There was no considerable increase in the photon fluence spectra from 6 to 15 MeV at the optimum thickness between 0.8 mm and 2 mm of tungsten. The optimum dimensions of the collimator were determined to be a length of 140 mm with an aperture of 5 mm  ×  70 mm for iron in a slit shape. According to the neutron yield, the best thickness obtained for the studied materials was 30 mm. The number of neutrons generated in BeO achieved the maximum value at 6 MeV, unlike that in Be, where the highest number of neutrons was observed at 15 MeV. Statistical uncertainty in all simulations was less than 0.3% and 0.05% for MCNPX and the standard electromagnetic (EM) physics packages of Geant4, respectively. Differences among spectra in various regions are due to various cross-section and stopping power data and different simulations of the physics processes.

  14. Estimation of photoneutron yield in linear accelerator with different collimation systems by Geant4 and MCNPX simulation codes.

    PubMed

    Kim, Yoon Sang; Khazaei, Zeinab; Ko, Junho; Afarideh, Hossein; Ghergherehchi, Mitra

    2016-04-01

    At present, the bremsstrahlung photon beams produced by linear accelerators are the most commonly employed method of radiotherapy for tumor treatments. A photoneutron source based on three different energies (6, 10 and 15 MeV) of a linac electron beam was designed by means of Geant4 and Monte Carlo N-Particle eXtended (MCNPX) simulation codes. To obtain maximum neutron yield, two arrangements for the photo neutron convertor were studied: (a) without a collimator, and (b) placement of the convertor after the collimator. The maximum photon intensities in tungsten were 0.73, 1.24 and 2.07 photon/e at 6, 10 and 15 MeV, respectively. There was no considerable increase in the photon fluence spectra from 6 to 15 MeV at the optimum thickness between 0.8 mm and 2 mm of tungsten. The optimum dimensions of the collimator were determined to be a length of 140 mm with an aperture of 5 mm  ×  70 mm for iron in a slit shape. According to the neutron yield, the best thickness obtained for the studied materials was 30 mm. The number of neutrons generated in BeO achieved the maximum value at 6 MeV, unlike that in Be, where the highest number of neutrons was observed at 15 MeV. Statistical uncertainty in all simulations was less than 0.3% and 0.05% for MCNPX and the standard electromagnetic (EM) physics packages of Geant4, respectively. Differences among spectra in various regions are due to various cross-section and stopping power data and different simulations of the physics processes.

  15. Estimation of photoneutron yield in linear accelerator with different collimation systems by Geant4 and MCNPX simulation codes.

    PubMed

    Kim, Yoon Sang; Khazaei, Zeinab; Ko, Junho; Afarideh, Hossein; Ghergherehchi, Mitra

    2016-04-01

    At present, the bremsstrahlung photon beams produced by linear accelerators are the most commonly employed method of radiotherapy for tumor treatments. A photoneutron source based on three different energies (6, 10 and 15 MeV) of a linac electron beam was designed by means of Geant4 and Monte Carlo N-Particle eXtended (MCNPX) simulation codes. To obtain maximum neutron yield, two arrangements for the photo neutron convertor were studied: (a) without a collimator, and (b) placement of the convertor after the collimator. The maximum photon intensities in tungsten were 0.73, 1.24 and 2.07 photon/e at 6, 10 and 15 MeV, respectively. There was no considerable increase in the photon fluence spectra from 6 to 15 MeV at the optimum thickness between 0.8 mm and 2 mm of tungsten. The optimum dimensions of the collimator were determined to be a length of 140 mm with an aperture of 5 mm  ×  70 mm for iron in a slit shape. According to the neutron yield, the best thickness obtained for the studied materials was 30 mm. The number of neutrons generated in BeO achieved the maximum value at 6 MeV, unlike that in Be, where the highest number of neutrons was observed at 15 MeV. Statistical uncertainty in all simulations was less than 0.3% and 0.05% for MCNPX and the standard electromagnetic (EM) physics packages of Geant4, respectively. Differences among spectra in various regions are due to various cross-section and stopping power data and different simulations of the physics processes. PMID:26975304

  16. A four mirror anastigmat collimator design for optical payload calibration

    NASA Astrophysics Data System (ADS)

    Rolt, Stephen; Calcines, Ariadna; Lomanowski, Bart A.; Bramall, David G.

    2016-07-01

    We present here a four mirror anastigmatic optical collimator design intended for the calibration of an earth observation satellite instrument. Specifically, the collimator is to be applied to the ground based calibration of the Sentinel-4/UVN instrument. This imaging spectrometer instrument itself is expected to be deployed in 2019 in a geostationary orbit and will make spatially resolved spectroscopic measurements of atmospheric contaminants. The collimator is to be deployed during the ground based calibration only and does not form part of the instrument itself. The purpose of the collimator is to provide collimated light within the two instrument passbands in the UV-VIS (305 - 500 nm) and the NIR (750 - 775 nm). Moreover, that collimated light will be derived from a variety of slit like objects located at the input focal (object) plane of the collimator which is uniformly illuminated by a number of light sources. The collimator must relay these objects with exceptionally high fidelity. To this end, the wavefront error of the collimator should be less than 30 nm rms across the collimator field of view. This field is determined by the largest object which is a large rectangular slit, 4.4° x 0.25°. Other important considerations affecting the optical design are the requirements for input telecentricity and the size (85 mm) and location (2500 mm `back focal distance') of the exit pupil. The design of the instrument against these basic requirements is discussed in detail. In addition an analysis of the straylight and tolerancing is presented in detail.

  17. The HEAO-1 Scanning Modulation Collimator

    NASA Astrophysics Data System (ADS)

    Schwartz, Daniel A.

    2013-01-01

    My niche on this panel seems to be the High Energy Astronomy Observatory-1 Scanning Modulation Collimator experiment. Our chair, Hale Bradt, and the late Herb Gursky each proposed a different version modulation collimator, which was condensed by NASA via "forced marriage," to the SMC. I worked as Project Scientist under Herb, later inheriting the PI role. The MIT Project Scientist, the late Rodger Doxsey, and I were told "this is your experiment," and "we are a seamless team regardless of institution." Rodger and I were young enough to believe this, and we made it happen (and not always with the best results vis a vis higher internal management). I was never interested in astronomy, and allegedly am still not. Why do an astro-metrical job of measuring and reporting the coordinates of X-ray sources? In fact we participated widely in the identification of the sources with astronomical object, and making each paper a discussion of the physics of the emission. An enjoyable way to learn some astronomy. The stated purpose of the Gursky/Bradt experiment was to enable optical identifications so that more detailed study could be done. I remember meeting with John Whelan to discuss his collaboration in making the optical identifications. He said he only wanted to study sources after they were identified. For many milliseconds I became very angry - "who is going to to the work to MAKE those identifications," but luckily before speaking I realized how satisfying it was that astronomers indeed wanted to study X-ray sources in other wavebands. The second biggest excitement in the HEAO-1 program was the "glitches" that appeared in the gyro data during final functional testing. This took some high-powered politics by all the PI's to convince MSFC to delay for 4 months, replacing the "funny" unit with one from HEAO-2 (Einstein) and later refurbishing that unit. Third biggest excitement was when a computer failed and final checkout during countdown at the Cape was done by looking at

  18. High resolution collimator system for X-ray detector

    DOEpatents

    Eberhard, Jeffrey W.; Cain, Dallas E.

    1987-01-01

    High resolution in an X-ray computerized tomography (CT) inspection system is achieved by using a collimator/detector combination to limit the beam width of the X-ray beam incident on a detector element to the desired resolution width. In a detector such as a high pressure Xenon detector array, a narrow tapered collimator is provided above a wide detector element. The collimator slits have any desired width, as small as a few mils at the top, the slit width is easily controlled, and they are fabricated on standard machines. The slit length determines the slice thickness of the CT image.

  19. Hollow Electron Beam Collimator: R and D Status Report

    SciTech Connect

    Stancari, G.; Drozhdin, A.; Kuznetsov, G.; Shiltsev, V.; Valishev, A.; Vorobiev, L.; Kabantsev, A.

    2010-11-04

    Magnetically confined hollow electron beams for controlled halo removal in high-energy colliders such as the Tevatron or the LHC may extend traditional collimation systems beyond the intensity limits imposed by tolerable material damage. They may also improve collimation performance by suppressing loss spikes due to beam jitter and by increasing capture efficiency. A hollow electron gun was designed and built. Its performance and stability were measured at the Fermilab test stand. The gun will be installed in one of the existing Tevatron electron lenses for preliminary tests of the hollow-beam collimator concept, addressing critical issues such as alignment and instabilities of the overlapping proton and electron beams.

  20. Development of hollow electron beams for proton and ion collimation

    SciTech Connect

    Stancari, G.; Drozhdin, A.I.; Kuznetsov, G.; Shiltsev, V.; Still, D.A.; Valishev, A.; Vorobiev, L.G.; Assmann, R.; Kabantsev, A.; /UC, San Diego

    2010-06-01

    Magnetically confined hollow electron beams for controlled halo removal in high-energy colliders such as the Tevatron or the LHC may extend traditional collimation systems beyond the intensity limits imposed by tolerable material damage. They may also improve collimation performance by suppressing loss spikes due to beam jitter and by increasing capture efficiency. A hollow electron gun was designed and built. Its performance and stability were measured at the Fermilab test stand. The gun will be installed in one of the existing Tevatron electron lenses for preliminary tests of the hollow-beam collimator concept, addressing critical issues such as alignment and instabilities of the overlapping proton and electron beams.

  1. LED collimation module for time sequential back light system

    NASA Astrophysics Data System (ADS)

    Lin, Jou-Hui; Lee, Chi-Hung; Tai, An-Ching; Chen, Yu-Hsuan; Ma, Shih-Hsin

    2014-09-01

    In this study, a collimated LED light source has been developed as a colorful liquid crystal display backlight, which is driven by a two-field driving scheme to display color. In each field, the angular rays of two colors from LEDs are collimated by a collimation lens, redirected by a light guide and converged by a cylindrical-lens array to map into corresponding sub-pixel positions for efficiently displaying color image. The simulation results of the backlight module and the corresponding experimental results will be discussed in detail.

  2. Tracking studies of the Compact Linear Collider collimation system

    SciTech Connect

    Agapov, I.; Burkhardt, H.; Schulte, D.; Latina, A.; Blair, G.A.; Malton, S.; Resta-Lopez, J.; /Oxford U., JAI

    2009-08-01

    A collimation system performance study includes several types of computations performed by different codes. Optics calculations are performed with codes such as MADX, tracking studies including additional effects such as wakefields, halo and tail generation, and dynamical machine alignment are done with codes such as PLACET, and energy deposition can be studied with BDSIM. More detailed studies of hadron production in the beam halo interaction with collimators are better performed with GEANT4 and FLUKA. A procedure has been developed that allows one to perform a single tracking study using several codes simultaneously. In this paper we study the performance of the Compact Linear Collider collimation system using such a procedure.

  3. Momentum slits, collimators and masks in the SLC

    SciTech Connect

    Walz, D.R.; McFarlane, A.; Lewandowski, E.; Zabdyr, J.

    1989-04-01

    The high specific power densities in the SLC give rise to a multitude of challenging problems in collimation and momentum analysis, beam containment, machine protection and background control. The results of an extensive program to develop most of the devices deemed necessary for operation of the arcs matching sections and the final focus region are presented. Emphasis is placed on materials selection and on unique features of remotely adjustable slits and halo clipper collimators which have to operate with great precision in a high-radiation, ultra-high vacuum environment. Also covered are solutions for a few fixed aperture machine protection collimators. 7 refs., 5 figs.

  4. Photon-photon collisions

    SciTech Connect

    Brodsky, S.J.

    1985-01-01

    The study of photon-photon collisions has progressed enormously, stimulated by new data and new calculational tools for QCD. In the future we can expect precise determinations of ..cap alpha../sub s/ and ..lambda../sup ms/ from the ..gamma..*..gamma.. ..-->.. ..pi../sup 0/ form factor and the photon structure function, as well as detailed checks of QCD, determination of the shape of the hadron distribution amplitudes from ..gamma gamma.. ..-->.. H anti H, reconstruction of sigma/sub ..gamma gamma../ from exclusive channels at low W/sub ..gamma gamma../, definitive studies of high p/sub T/ hadron and jet production, and studies of threshold production of charmed systems. Photon-photon collisions, along with radiative decays of the psi and UPSILON, are ideal for the study of multiquark and gluonic resonances. We have emphasized the potential for resonance formation near threshold in virtually every hadronic exclusive channel, including heavy quark states c anti c c anti c, c anti c u anti u, etc. At higher energies SLC, LEP, ...) parity-violating electroweak effects and Higgs production due to equivalent Z/sup 0/ and W/sup + -/ beams from e ..-->.. eZ/sup 0/ and e ..-->.. nu W will become important. 44 references.

  5. Compton backscattered collimated x-ray source

    DOEpatents

    Ruth, R.D.; Huang, Z.

    1998-10-20

    A high-intensity, inexpensive and collimated x-ray source is disclosed for applications such as x-ray lithography is disclosed. An intense pulse from a high power laser, stored in a high-finesse resonator, repetitively collides nearly head-on with and Compton backscatters off a bunched electron beam, having relatively low energy and circulating in a compact storage ring. Both the laser and the electron beams are tightly focused and matched at the interaction region inside the optical resonator. The laser-electron interaction not only gives rise to x-rays at the desired wavelength, but also cools and stabilizes the electrons against intrabeam scattering and Coulomb repulsion with each other in the storage ring. This cooling provides a compact, intense bunch of electrons suitable for many applications. In particular, a sufficient amount of x-rays can be generated by this device to make it an excellent and flexible Compton backscattered x-ray (CBX) source for high throughput x-ray lithography and many other applications. 4 figs.

  6. Compton backscattered collimated x-ray source

    DOEpatents

    Ruth, Ronald D.; Huang, Zhirong

    1998-01-01

    A high-intensity, inexpensive and collimated x-ray source for applications such as x-ray lithography is disclosed. An intense pulse from a high power laser, stored in a high-finesse resonator, repetitively collides nearly head-on with and Compton backscatters off a bunched electron beam, having relatively low energy and circulating in a compact storage ring. Both the laser and the electron beams are tightly focused and matched at the interaction region inside the optical resonator. The laser-electron interaction not only gives rise to x-rays at the desired wavelength, but also cools and stabilizes the electrons against intrabeam scattering and Coulomb repulsion with each other in the storage ring. This cooling provides a compact, intense bunch of electrons suitable for many applications. In particular, a sufficient amount of x-rays can be generated by this device to make it an excellent and flexible Compton backscattered x-ray (CBX) source for high throughput x-ray lithography and many other applications.

  7. Performance of a collimating L-shaped laterally graded multilayer mirror for the IXS analyzer system at NSLS-II.

    PubMed

    Suvorov, Alexey; Coburn, David S; Cunsolo, Alessandro; Keister, Jeffrey W; Upton, Mary H; Cai, Yong Q

    2014-05-01

    The L-shaped laterally graded multilayer mirror is a vital part of the ultrahigh-energy and momentum-resolution inelastic X-ray scattering spectrometer at the National Synchrotron Light Source II. This mirror was designed and implemented as a two-dimensional collimating optic for the analyzer system. Its performance was characterized using a secondary large-divergence source at the 30-ID beamline of the Advanced Photon Source, which yielded an integrated reflectivity of 47% and a collimated beam divergence of 78 µrad with a source size of 10 µm. Numerical simulations of the mirror performance in tandem with the analyzer crystal optics provided details on the acceptance sample volume in forward scattering and defined the technical requirements on the mirror stability and positioning precision. It was shown that the mirror spatial and angular stability must be in the range <8.4 µm and <21.4 µrad, respectively, for reliable operation of the analyzer.

  8. Bench-Top Impedance Measurements for a Rotatable Copper Collimator for the LHC Phase II Collimation Upgrade

    SciTech Connect

    Smith, Jeffrey Claiborne; Bane, Karl; Doyle, Eric; Keller, Lew; Lundgren, Steve; Markiewicz, Tom; Ng, Cho-Kuen; Xiao, Liling; /SLAC

    2010-08-26

    Simulations have been performed in Omega3P to study both trapped modes and impedance contributions of a rotatable collimator for the LHC phase II collimation upgrade. Bench-top stretched coil probe impedance methods are also being implemented for measurements on prototype components to directly measure the low frequency impedance contributions. The collimator design also calls for a RF contact interface at both jaw ends with contact resistance much less than a milliohm in order to limit transverse impedance. DC resistance measurements in a custom built test chamber have been performed to test the performance of this interface.

  9. Comparison of the TESLA, NLC and CLIC beam collimation performance

    SciTech Connect

    Alexandr I Drozhdin et al.

    2003-03-27

    This note describes studies performed in the framework of the Collimation Task Force organized to support the work of the International Linear Collider Technical Review Committee. The post-linac beam-collimation systems in the TESLA, JLC/NLC and CLIC linear-collider designs are compared using the same computer code under the same assumptions. Their performance is quantified in terms of beam-halo and synchrotron-radiation collimation efficiency. The performance of the current designs varies across projects, and does not always meet the original design goals. But these comparisons suggest that achieving the required performance in a future linear collider is feasible. The post-TRC plans of the Collimation Task Force are briefly outlined in closing.

  10. Variable-energy collimator for high-energy radiation

    DOEpatents

    Hill, R.A.

    1982-03-03

    An apparatus is disclosed providing a variable aperture energy beam collimator. A plurality of beam opaque blocks are in sliding interface edge contact to form a variable aperture. The blocks may be offset at the apex angle to provide a non-equilateral aperture. A plurality of collimator block assemblies may be employed for providing a channel defining a collimated beam. Adjacent assemblies are inverted front-to-back with respect to one another for preventing noncollimated ;energy from emergine from the apparatus. An adjustment mechanism comprises a cable attached to at least one block and a hand wheel mechanism for operating the cable. The blocks are supported by guide rods engaging slide brackets on the blocks. The guide rods are pivotally connected at each end to intermediate actuators supported on rotatable shafts to change the shape of the aperture. A divergent collimated beam may be obtained by adjusting the apertures of adjacent stages to be unequal.

  11. Collimator design for the J-PEX sounding rocket

    NASA Astrophysics Data System (ADS)

    Golembiewski, Cara M.; Cruddace, Raymond G.; Kowalski, Michael P.

    1998-11-01

    The Joint Astrophysical Plasmadynamic Experiment is a high- resolution extreme UV spectrometer which operates at near- normal incidence and incorporates the most recent developments in reflecting multilayer gratings. The spectrometer comprises the following principal elements: collimators which define the 1.2 degrees field of view, a spherical multilayer-coated grating which consists of 4 segments, each 8 cm by 16 cm in area and with a focal length of 2.2 m, and a microchannel plate imaging detector. The collimator is an adaptation of a previous design which was modified to minimize the input of scattered airglow radiation while maintaining the maximum effective area within the desired field of view. This is achieved by blackening the collimator surfaces with an EBONOL-C process. In this paper we will discuss the design requirements and manufacturing techniques for the collimator.

  12. Slant-hole collimator, dual mode sterotactic localization method

    DOEpatents

    Weisenberger, Andrew G.

    2002-01-01

    The use of a slant-hole collimator in the gamma camera of dual mode stereotactic localization apparatus allows the acquisition of a stereo pair of scintimammographic images without repositioning of the gamma camera between image acquisitions.

  13. Electron Beam Collimation for the Next Generation Light Source

    SciTech Connect

    Steier, C.; Emma, P.; Nishimura, H.; Papadopoulos, C.; Sannibale, F.

    2013-05-20

    The Next Generation Light Source will deliver high (MHz) repetition rate electron beams to an array of free electron lasers. Because of the significant average current in such a facility, effective beam collimation is extremely important to minimize radiation damage to undulators, prevent quenches of superconducting cavities, limit dose rates outside of the accelerator tunnel and prevent equipment damage. This paper describes the early conceptual design of a collimation system, as well as initial results of simulations to test its effectiveness.

  14. Beam loss by collimation in a neutralizer duct

    SciTech Connect

    Hamilton, G.W.; Willmann, P.A.

    1980-04-03

    Beam fractions lost by collimation in a neutralizer duct are computed in x-x' phase space by using three examples of slab beam distributions under a broad range of duct dimensions, beam half-widths, and beam divergences. The results can be used to design compact neutralizers and to specify beam requirements. The computer code ILOST can be used under a broad range of beam conditions to compute the fraction lost by collimation.

  15. Steering and collimating ballistic electrons with amphoteric refraction

    SciTech Connect

    Radu, A.; Dragoman, D.; Iftimie, S.

    2012-07-15

    We show that amphoteric refraction of ballistic electrons, i.e., positive or negative refraction depending on the incidence angle, occurs at an interface between an isotropic and an anisotropic medium and can be employed to steer and collimate electron beams. The steering angle is determined by the materials' parameters, but the degree of collimation can be tuned in a significant range by changing the energy of ballistic electrons.

  16. Benchmarking of collimation tracking using RHIC beam loss data.

    SciTech Connect

    Robert-Demolaize,G.; Drees, A.

    2008-06-23

    State-of-the-art tracking tools were recently developed at CERN to study the cleaning efficiency of the Large Hadron Collider (LHC) collimation system. In order to estimate the prediction accuracy of these tools, benchmarking studies can be performed using actual beam loss measurements from a machine that already uses a similar multistage collimation system. This paper reviews the main results from benchmarking studies performed with specific data collected from operations at the Relativistic Heavy Ion Collider (RHIC).

  17. Parallel-hole collimator concept for stationary SPECT imaging

    NASA Astrophysics Data System (ADS)

    Pato, Lara R. V.; Vandenberghe, Stefaan; Zedda, Tiziana; Van Holen, Roel

    2015-11-01

    Parallel-hole SPECT collimators have traditionally been manufactured by stacking sheets of lead foil or by casting. These techniques significantly restrict our options in terms of collimator geometry. However, recent developments in metal additive manufacturing are making novel collimator designs possible, giving rise to new opportunities in SPECT imaging. In this paper we propose an innovative type of collimator for stationary SPECT, using parallel-holes whose collimation direction depends on their axial position. Its main advantage compared to current stationary SPECT systems (which are based on pinholes) is that, using only axial bed translations, we can achieve complete angular sampling of an increased portion of the transaxial area of the collimator bore. This allows the system to be much more compact than current stationary SPECT systems that image objects of the same size. We describe three possible designs, for full-body, brain and small-animal imaging, respectively, and test their feasibility using simulations. The system modeling method is validated against realistic Monte Carlo simulations, and then used in the evaluation of the systems’ performances and reconstructions. The simulations show that the system is able to reconstruct objects occupying the predicted field of view (75% of the transaxial area of the bore) without sampling artifacts. In particular, we perform reconstructions from noisy projection data obtained for an activity and scanning time similar to standard protocols for the three applications, and the resulting images indicate the possibility of using the proposed systems in practice.

  18. Measurements of the Transverse Wakefields Due to Varying Collimator Characteristics

    SciTech Connect

    Molloy, S.; Seletskiy, Sergei; Woods, Mike; Smith, Jonathan David Andrew; Beard, Carl David; Fernandez-Hernando, Juan Luis; Watson, Nigel; Bungau, Adriana; Sopczak, Andre; /Lancaster U.

    2007-07-06

    We report on measurements of the transverse wakefields induced by collimators of differing characteristics. An apparatus allowing the insertion of different collimator jaws into the path of a beam was installed in End Station A (ESA) in SLAC. Eight comparable collimator geometries were designed, including one that would allow easy comparison with previous results, and were installed in this apparatus. Measurements of the beam kick due to the collimator wakefields were made with a beam energy of 28.5 GeV, and beam dimensions of 100 microns vertically and a range of 0.5 to 1.5 mm longitudinally. The trajectory of the beam upstream and downstream of the collimator test apparatus was determined from the outputs of ten BPMs (four upstream and six downstream), thus allowing a measurement of the angular kick imparted to the beam by the collimator under test. The transverse wakefield was inferred from the measured kick. The different aperture designs, data collection and analysis, and initial comparison to theoretical and analytic predictions are presented here.

  19. Dual self-image technique for beam collimation

    NASA Astrophysics Data System (ADS)

    Herrera-Fernandez, Jose Maria; Sanchez-Brea, Luis Miguel; Torcal-Milla, Francisco Jose; Morlanes, Tomas; Bernabeu, Eusebio

    2016-07-01

    We propose an accurate technique for obtaining highly collimated beams, which also allows testing the collimation degree of a beam. It is based on comparing the period of two different self-images produced by a single diffraction grating. In this way, variations in the period of the diffraction grating do not affect to the measuring procedure. Self-images are acquired by two CMOS cameras and their periods are determined by fitting the variogram function of the self-images to a cosine function with polynomial envelopes. This way, loss of accuracy caused by imperfections of the measured self-images is avoided. As usual, collimation is obtained by displacing the collimation element with respect to the source along the optical axis. When the period of both self-images coincides, collimation is achieved. With this method neither a strict control of the period of the diffraction grating nor a transverse displacement, required in other techniques, are necessary. As an example, a LED considering paraxial approximation and point source illumination is collimated resulting a resolution in the divergence of the beam of δ φ =+/- 1.57 μ {rad}.

  20. A clinical gamma camera-based pinhole collimated system for high resolution small animal SPECT imaging.

    PubMed

    Mejia, J; Galvis-Alonso, O Y; Castro, A A de; Braga, J; Leite, J P; Simões, M V

    2010-12-01

    The main objective of the present study was to upgrade a clinical gamma camera to obtain high resolution tomographic images of small animal organs. The system is based on a clinical gamma camera to which we have adapted a special-purpose pinhole collimator and a device for positioning and rotating the target based on a computer-controlled step motor. We developed a software tool to reconstruct the target's three-dimensional distribution of emission from a set of planar projections, based on the maximum likelihood algorithm. We present details on the hardware and software implementation. We imaged phantoms and heart and kidneys of rats. When using pinhole collimators, the spatial resolution and sensitivity of the imaging system depend on parameters such as the detector-to-collimator and detector-to-target distances and pinhole diameter. In this study, we reached an object voxel size of 0.6 mm and spatial resolution better than 2.4 and 1.7 mm full width at half maximum when 1.5- and 1.0-mm diameter pinholes were used, respectively. Appropriate sensitivity to study the target of interest was attained in both cases. Additionally, we show that as few as 12 projections are sufficient to attain good quality reconstructions, a result that implies a significant reduction of acquisition time and opens the possibility for radiotracer dynamic studies. In conclusion, a high resolution single photon emission computed tomography (SPECT) system was developed using a commercial clinical gamma camera, allowing the acquisition of detailed volumetric images of small animal organs. This type of system has important implications for research areas such as Cardiology, Neurology or Oncology.

  1. EDITORIAL: Photonic Crystal Devices

    NASA Astrophysics Data System (ADS)

    Bhattacharya, Pallab K.

    2007-05-01

    glimpse of some of the most recent advances in the application of photonic crystals. The modelling of PC defect-mode cavities are described by Zhou et al. Ye and co-authors describe the concept and realization of a novel 3D silicon-based spiral PC. It is, in fact, the only article on 3D PCs. The design and realization of ultra-high Q heterostructure PC nanocavities are described by Song and co-authors. The concept of self-collimation of light in PCs and its applications are presented by Prather and co-workers. Experimental and numerical studies on the negative refraction related phenomenon in 2D PCs are the subject of the next article by Ozbay and co-authors. The emerging subject of slow light generation, control and propagation in PCs is presented in the next two articles by Baba and Mori and by Krauss. Finally, the progress made in the development of PC microcavity lasers and electrically injected microcavity light emitters and arrays is described, respectively, by O'Brien et al and by Chakravarty et al. It is hoped that readers will get a sense of the exciting developments and the possibilities presented by heterostructure photonic crystals and their devices from reading the articles in this Cluster.

  2. Parameterized source term in the diffusion approximation for enhanced near-field modeling of collimated light

    NASA Astrophysics Data System (ADS)

    Jia, Mengyu; Wang, Shuang; Chen, Xueying; Gao, Feng; Zhao, Huijuan

    2016-03-01

    Most analytical methods for describing light propagation in turbid medium exhibit low effectiveness in the near-field of a collimated source. Motivated by the Charge Simulation Method in electromagnetic theory as well as the established discrete source based modeling, we have reported on an improved explicit model, referred to as "Virtual Source" (VS) diffuse approximation (DA), to inherit the mathematical simplicity of the DA while considerably extend its validity in modeling the near-field photon migration in low-albedo medium. In this model, the collimated light in the standard DA is analogously approximated as multiple isotropic point sources (VS) distributed along the incident direction. For performance enhancement, a fitting procedure between the calculated and realistic reflectances is adopted in the nearfield to optimize the VS parameters (intensities and locations). To be practically applicable, an explicit 2VS-DA model is established based on close-form derivations of the VS parameters for the typical ranges of the optical parameters. The proposed VS-DA model is validated by comparing with the Monte Carlo simulations, and further introduced in the image reconstruction of the Laminar Optical Tomography system.

  3. Planar integrated metasurfaces for highly-collimated terahertz quantum cascade lasers.

    PubMed

    Liang, Guozhen; Dupont, Emmanuel; Fathololoumi, Saeed; Wasilewski, Zbigniew R; Ban, Dayan; Liang, Hou Kun; Zhang, Ying; Yu, Siu Fung; Li, Lianhe H; Davies, Alexander Giles; Linfield, Edmund H; Liu, Hui Chun; Wang, Qi Jie

    2014-11-18

    We report planar integration of tapered terahertz (THz) frequency quantum cascade lasers (QCLs) with metasurface waveguides that are designed to be spoof surface plasmon (SSP) out-couplers by introducing periodically arranged SSP scatterers. The resulting surface-emitting THz beam profile is highly collimated with a divergence as narrow as ~4° × 10°, which indicates a good waveguiding property of the metasurface waveguide. In addition, the low background THz power implies a high coupling efficiency for the THz radiation from the laser cavity to the metasurface structure. Furthermore, since all the structures are in-plane, this scheme provides a promising platform where well-established surface plasmon/metasurface techniques can be employed to engineer the emitted beam of THz QCLs controllably and flexibly. More importantly, an integrated active THz photonic circuit for sensing and communication applications could be constructed by incorporating other optoelectronic devices such as Schottky diode THz mixers, and graphene modulators and photodetectors.

  4. Development of advanced industrial SPECT system with 12-gonal diverging-collimator.

    PubMed

    Park, Jang Guen; Jung, Sung-Hee; Kim, Jong Bum; Moon, Jinho; Han, Min Cheol; Kim, Chan Hyeong

    2014-07-01

    Industrial single photon emission computed tomography (SPECT) is a promising diagnosis technique to investigate the dynamic behavior of process media. In the present study, a 12-gonal industrial SPECT system was developed using diverging collimators, and its performance was compared with those of hexagonal and 24-gonal systems. Of all of the systems, the 12-gonal type showed the best performance, providing (1) a detection-efficiency map without edge artifacts, (2) the best image resolution, and (3) reconstruction images that correctly furnish multi-source information. Based on the performance of the three different types of configurations, a SPECT system with 12-gonal type configuration was found most suitable for investigating and visualization of flow dynamics in industrial process systems.

  5. Development of advanced industrial SPECT system with 12-gonal diverging-collimator.

    PubMed

    Park, Jang Guen; Jung, Sung-Hee; Kim, Jong Bum; Moon, Jinho; Han, Min Cheol; Kim, Chan Hyeong

    2014-07-01

    Industrial single photon emission computed tomography (SPECT) is a promising diagnosis technique to investigate the dynamic behavior of process media. In the present study, a 12-gonal industrial SPECT system was developed using diverging collimators, and its performance was compared with those of hexagonal and 24-gonal systems. Of all of the systems, the 12-gonal type showed the best performance, providing (1) a detection-efficiency map without edge artifacts, (2) the best image resolution, and (3) reconstruction images that correctly furnish multi-source information. Based on the performance of the three different types of configurations, a SPECT system with 12-gonal type configuration was found most suitable for investigating and visualization of flow dynamics in industrial process systems. PMID:24636865

  6. Transverse angular momentum of photons

    SciTech Connect

    Aiello, Andrea

    2010-05-15

    We develop the quantum theory of transverse angular momentum of light beams. The theory applies to paraxial and quasiparaxial photon beams in vacuum and reproduces the known results for classical beams when applied to coherent states of the field. Both the Poynting vector, alias the linear momentum, and the angular-momentum quantum operators of a light beam are calculated including contributions from first-order transverse derivatives. This permits a correct description of the energy flow in the beam and the natural emergence of both the spin and the angular momentum of the photons. We show that for collimated beams of light, orbital angular-momentum operators do not satisfy the standard commutation rules. Finally, we discuss the application of our theory to some concrete cases.

  7. Correction for collimator-detector response in SPECT using point spread function template.

    PubMed

    Chun, Se Young; Fessler, Jeffrey A; Dewaraja, Yuni K

    2013-02-01

    Compensating for the collimator-detector response (CDR) in SPECT is important for accurate quantification. The CDR consists of both a geometric response and a septal penetration and collimator scatter response. The geometric response can be modeled analytically and is often used for modeling the whole CDR if the geometric response dominates. However, for radionuclides that emit medium or high-energy photons such as I-131, the septal penetration and collimator scatter response is significant and its modeling in the CDR correction is important for accurate quantification. There are two main methods for modeling the depth-dependent CDR so as to include both the geometric response and the septal penetration and collimator scatter response. One is to fit a Gaussian plus exponential function that is rotationally invariant to the measured point source response at several source-detector distances. However, a rotationally-invariant exponential function cannot represent the star-shaped septal penetration tails in detail. Another is to perform Monte-Carlo (MC) simulations to generate the depth-dependent point spread functions (PSFs) for all necessary distances. However, MC simulations, which require careful modeling of the SPECT detector components, can be challenging and accurate results may not be available for all of the different SPECT scanners in clinics. In this paper, we propose an alternative approach to CDR modeling. We use a Gaussian function plus a 2-D B-spline PSF template and fit the model to measurements of an I-131 point source at several distances. The proposed PSF-template-based approach is nearly non-parametric, captures the characteristics of the septal penetration tails, and minimizes the difference between the fitted and measured CDR at the distances of interest. The new model is applied to I-131 SPECT reconstructions of experimental phantom measurements, a patient study, and a MC patient simulation study employing the XCAT phantom. The proposed model

  8. Multiple pinhole collimator based microscopic x-ray luminescence computed tomography

    NASA Astrophysics Data System (ADS)

    Zhang, Wei; Zhu, Dianwen; Li, Changqing

    2016-03-01

    X-ray luminescence computed tomography (XLCT) is a new hybrid imaging modality, which has the capability to improve optical spatial resolution to hundreds of micrometers for deep targets. In this paper, we report a multiple pinhole collimator based microscopic X-ray luminescence computed tomography (microXLCT) system for small animal imaging. Superfine collimated X-ray pencil beams are used to excite deeply embedded phosphor particles, allowing us to obtain sub-millimeter optical spatial resolution in deep tissues. Multiple collimated X-ray beams are generated by mounting an array of pinholes in the front of a powerful X-ray tube. With multiple X-ray beams scanning, the phosphor particles in the region of the multiple beams are excited simultaneously, which requires less scanning time compared with a single beam scanning. The emitted optical photons on the top surface of the phantom are measured with an electron multiplying charge-coupled device (EMCCD) camera. Meanwhile, an X-ray detector is used to determine the X-ray beam size and position, which are used as structural guidance in the microXLCT image reconstruction. To validate the performance of our proposed multiple pinhole based microXLCT imaging system, we have performed numerical simulations and a phantom experiment. In the numerical simulations, we simulated a cylindrical phantom with two and six embedded targets, respectively. In the simulations, we used four parallel X-ray beams with the beam diameter of 0.1 mm and the beam interval of 3.2 mm. We can reconstruct deeply embedded multiple targets with a target diameter of 0.2 mm using measurements in six projections, which indicated that four parallel X-ray beam scan could reduce scanning time without comprising the reconstructed image quality. In the phantom experiment, we generated two parallel X-ray beams with the beam diameter of 0.5 mm and the beam interval of 4.2 mm. We scanned a phantom of one target with the two parallel X-ray beams. The target was

  9. A magnetically collimated jet from an evolved star.

    PubMed

    Vlemmings, Wouter H T; Diamond, Philip J; Imai, Hiroshi

    2006-03-01

    Planetary nebulae often have asymmetric shapes, even though their progenitor stars were symmetric; this structure could be the result of collimated jets from the evolved stars before they enter the planetary nebula phase. Theoretical models have shown that magnetic fields could be the dominant source of jet-collimation in evolved stars, just as these fields are thought to collimate outflows in other astrophysical sources, such as active galactic nuclei and proto-stars. But hitherto there have been no direct observations of both the magnetic field direction and strength in any collimated jet. Here we report measurements of the polarization of water vapour masers that trace the precessing jet emanating from the asymptotic giant branch star W43A (at a distance of 2.6 kpc from the Sun), which is undergoing rapid evolution into a planetary nebula. The masers occur in two clusters at opposing tips of the jets, approximately 1,000 au from the star. We conclude from the data that the magnetic field is indeed collimating the jet. PMID:16511488

  10. Radiation shielding for the Main Injector collimation system

    SciTech Connect

    Rakhno, Igor; /Fermilab

    2007-12-01

    The results of Monte Carlo radiation shielding studies performed with the MARS15 code for the Main Injector collimation system are presented and discussed. A Proton Plan was developed recently at Fermilab for the benefit of the existing neutrino programs as well as to increase anti-proton production for the Tevatron programs [1]. As a part of the plan, the intensity of proton beams in the Main Injector (MI) should be increased by means of slip-stacking injection. In order to localize beam loss associated with the injection, a collimation system was designed [2] that satisfies all the radiation and engineering constraints. The system itself comprises a primary collimator and four secondary ones to which various masks are added. It is assumed that beam loss at the slip-stacking injection is equal to 5% of total intensity which is 5.5 x 10{sup 13} protons per pulse [2]. As far as pulse separation is 2.2 seconds, one has (5.5 x 10{sup 13}/2.2) x 0.05 = 1.25 x 10{sup 12} protons interacting per second with the primary collimator. In the paper the geometry model of the corresponding MI region and beam loss model are described. The model of the region was built by means of the MAD-MARS Beam Line Builder (MMBLB) [3] using results of the collimation studies [2]. The results of radiation shielding calculations performed with the MARS15 code [4] are presented.

  11. Collimated fast electron beam generation in critical density plasma

    SciTech Connect

    Iwawaki, T. Habara, H.; Morita, K.; Tanaka, K. A.; Baton, S.; Fuchs, J.; Chen, S.; Nakatsutsumi, M.; Rousseaux, C.; Filippi, F.; Nazarov, W.

    2014-11-15

    Significantly collimated fast electron beam with a divergence angle 10° (FWHM) is observed when an ultra-intense laser pulse (I = 10{sup 14 }W/cm{sup 2}, 300 fs) irradiates a uniform critical density plasma. The uniform plasma is created through the ionization of an ultra-low density (5 mg/c.c.) plastic foam by X-ray burst from the interaction of intense laser (I = 10{sup 14 }W/cm{sup 2}, 600 ps) with a thin Cu foil. 2D Particle-In-Cell (PIC) simulation well reproduces the collimated electron beam with a strong magnetic field in the region of the laser pulse propagation. To understand the physical mechanism of the collimation, we calculate energetic electron motion in the magnetic field obtained from the 2D PIC simulation. As the results, the strong magnetic field (300 MG) collimates electrons with energy over a few MeV. This collimation mechanism may attract attention in many applications such as electron acceleration, electron microscope and fast ignition of laser fusion.

  12. Resistive collimation of electron beams in relativistic and degenerate plasma

    NASA Astrophysics Data System (ADS)

    Mahdavi, M.; Khodadadi Azadboni, F.

    2014-09-01

    The purpose of this research is the study of the effects of plasma state and fiber on collimating relativistic electron beam in fast ignition. In this paper, for collimating relativistic electrons produced at the laser plasma interaction, a thin fiber of aluminum, lithium or CH either in the classical, degenerate or relativistic plasma states is considered. The fast electron beam could be collimated down to radii of 10 μm, in that case, the best results are achieved when there is a sharp transition in resistance. This ensures that the correct magnetic growth rate is used for hot electrons at different energy levels. Calculations show that the resistivity of the material surrounding the CH fiber in the degenerate plasma is smaller than that for classical and relativistic plasma.

  13. Fermilab Main Injector Collimation Systems: Design, Commissioning and Operation

    SciTech Connect

    Brown, Bruce; Adamson, Philip; Capista, David; Drozhdin, A.I.; Johnson, David E.; Kourbanis, Ioanis; Mokhov, Nikolai V.; Morris, Denton K.; Rakhno, Igor; Seiya, Kiyomi; Sidorov, Vladimir; /Fermilab

    2009-05-01

    The Fermilab Main Injector is moving toward providing 400 kW of 120 GeV proton beams using slip stacking injection of eleven Booster batches. Loss of 5% of the beam at or near injection energy results in 1.5 kW of beam loss. A collimation system has been implemented to localize this loss with the design emphasis on beam not captured in the accelerating RF buckets. More than 95% of these losses are captured in the collimation region. We will report on the construction, commissioning and operation of this collimation system. Commissioning studies and loss measurement tools will be discussed. Residual radiation monitoring of the Main Injector machine components will be used to demonstrate the effectiveness of these efforts.

  14. Collimated neutron probe for soil water content measurements

    USGS Publications Warehouse

    Klenke, J.M.; Flint, A.L.

    1991-01-01

    A collimated neutron probe was designed to enable mesurements in specific directions from the access tube. To determine the size and shape of soil volume affecting the neutron counts, experiments were conducted to evaluate: 1) the vertical distance of soil above and below the probe that influences neutron counts; 2) the horizontal distance away from the probe into the soil that influences neutron counts; 3) the angle of soil viewed by the probe from the collimator; and 4) the three-dimensional thermal-neutron density field. The vertical distance was ~0.5m, the horizontal distance was ~0.2m, and the angle of soil viewed by the probe from the collimator was ~120??. Thermal neutrons detected from distances or angles larger than these values influence the determination of relative water content by 5% or less. -from Authors

  15. High energy collimating fine grids for HESP program

    NASA Technical Reports Server (NTRS)

    Eberhard, Carol D.; Frazier, Edward

    1993-01-01

    There is a need to develop fine pitch x-ray collimator grids as an enabling technology for planned future missions. The grids consist of an array of thin parallel strips of x-ray absorbing material, such as tungsten, with pitches ranging from 34 microns to 2.036 millimeters. The grids are the key components of a new class of spaceborne instruments known as 'x-ray modulation collimators.' These instruments are the first to produce images of celestial sources in the hard x-ray and gamma-ray spectral regions.

  16. ILC Beam delivery WG summary: Optics, collimation and background

    SciTech Connect

    Angal-Kalinin, D.; Jackson, F.; Mokhov, N.V.; Kuroda, S.; Seryi, A.A.; /SLAC

    2006-07-01

    The paper summarizes the work of the Beam Delivery working group (WG4) at Snowmass 2005 workshop, focusing on status of optics, layout, collimation and detector background. The strawman layout with two interaction regions was recommended at the first ILC workshop at KEK in November 2004. Two crossing-angle designs were included in this layout. The design of the ILC BDS has evolved since the first ILC workshop. The progress on the BDS design including the collimation system, and extraction line design have been reviewed and the design issues were discussed during the WG4 sessions at the Snowmass, and are described in this paper.

  17. Impedance Scaling for Small-angle Tapers and Collimators

    SciTech Connect

    Stupakov, G.; /SLAC

    2010-02-11

    In this note I will prove that the impedance calculated for a small-angle collimator or taper, of arbitrary 3D profile, has a scaling property that can greatly simplify numerical calculations. This proof is based on the parabolic equation approach to solving Maxwell's equation developed in Refs. [1, 2]. We start from the parabolic equation formulated in [3]. As discussed in [1], in general case this equation is valid for frequencies {omega} >> c/a where a is a characteristic dimension of the obstacle. However, for small-angle tapers and collimators, the region of validity of this equation extends toward smaller frequencies and includes {omega} {approx} c/a.

  18. Hydrodynamic collimation of gamma-ray-burst fireballs

    PubMed

    Levinson; Eichler

    2000-07-10

    Analytic solutions are presented for the hydrodynamic collimation of a relativistic fireball by a surrounding baryonic wind emanating from a torus. The opening angle is shown to be the ratio of the power output of the inner fireball to that of the exterior baryonic wind. The gamma ray burst 990123 might thus be interpreted as a baryon-poor jet (BPJ) with an energy output of order 10(50) erg or less, collimated by a baryonic wind from a torus with an energy output of order 10(52.5) erg, roughly the geometric mean of the BPJ and its isotropic equivalent.

  19. Measurements of Collimator Wakefields at End Station A

    SciTech Connect

    Fernandez-Hernando, J.L.; Molloy, S.; Smith, J.D.A.; Watson, Nigel Keith; /Birmingham U.

    2011-11-01

    The angular deflection of a 28.5 GeV electron beam passing off-axis between the jaws of a collimator, generating a transverse wakefield, were measured in End Station A (ESA) at SLAC. In total, fifteen different configurations of collimator geometry and material were tested: some were chosen for compatibility with previous measurements while others served to study the effect of geometry and taper angles (geometrical contribution to the wakefield) and the effect of the material resistivity (resistive contribution) to the imparted kick. This paper summarises the last update of preliminary experimental results before they are finalised. The reconstructed kick factor is compared to analytical calculations and simulations.

  20. Beamlet based direct aperture optimization for MERT using a photon MLC

    SciTech Connect

    Henzen, D. Manser, P.; Frei, D.; Volken, W.; Born, E. J.; Joosten, A.; Lössl, K.; Aebersold, D. M.; Chatelain, C.; Fix, M. K.; Neuenschwander, H.; Stampanoni, M. F. M.

    2014-12-15

    Purpose: A beamlet based direct aperture optimization (DAO) for modulated electron radiotherapy (MERT) using photon multileaf collimator (pMLC) shaped electron fields is developed and investigated. Methods: The Swiss Monte Carlo Plan (SMCP) allows the calculation of dose distributions for pMLC shaped electron beams. SMCP is interfaced with the Eclipse TPS (Varian Medical Systems, Palo Alto, CA) which can thus be included into the inverse treatment planning process for MERT. This process starts with the import of a CT-scan into Eclipse, the contouring of the target and the organs at risk (OARs), and the choice of the initial electron beam directions. For each electron beam, the number of apertures, their energy, and initial shape are defined. Furthermore, the DAO requires dose–volume constraints for the structures contoured. In order to carry out the DAO efficiently, the initial electron beams are divided into a grid of beamlets. For each of those, the dose distribution is precalculated using a modified electron beam model, resulting in a dose list for each beamlet and energy. Then the DAO is carried out, leading to a set of optimal apertures and corresponding weights. These optimal apertures are now converted into pMLC shaped segments and the dose calculation for each segment is performed. For these dose distributions, a weight optimization process is launched in order to minimize the differences between the dose distribution using the optimal apertures and the pMLC segments. Finally, a deliverable dose distribution for the MERT plan is obtained and loaded back into Eclipse for evaluation. For an idealized water phantom geometry, a MERT treatment plan is created and compared to the plan obtained using a previously developed forward planning strategy. Further, MERT treatment plans for three clinical situations (breast, chest wall, and parotid metastasis of a squamous cell skin carcinoma) are created using the developed inverse planning strategy. The MERT plans are

  1. Iterative reconstruction with correction of the spatially variant fan-beam collimator response in neurotransmission SPET imaging.

    PubMed

    Pareto, Deborah; Cot, Albert; Pavía, Javier; Falcón, Carles; Juvells, Ignacio; Lomeña, Francisco; Ros, Domènec

    2003-10-01

    The dopamine transporter (DAT) has been shown to be a sensitive indicator of nigrostriatal dopamine function. Although visual inspection is often sufficient to assess DAT imaging, quantification could improve the diagnostic accuracy of single-photon emission tomography (SPET) studies of the dopaminergic system. The aim of this study was to assess the accuracy of quantification of the striatal/background uptake ratio when correction for attenuation, scatter and spatially variant fan-beam collimator response is performed in technetium-99m and iodine-123 SPET imaging. A numerical striatal phantom was implemented, and simulated projections of low-energy photons were obtained by using the SimSET Monte Carlo code. High-energy contamination in 123I studies was modelled from experimental measurements with 99mTc and 123I. The ordered subsets expectation maximisation (OSEM) algorithm was employed in reconstruction. Mean improvements of 8% and 16% were obtained in the calculated striatal/background uptake ratio in the putamen and the caudate, respectively, when the spatially variant point spread function was included in the transition matrix. Ideal scatter correction resulted in improvements in the putamen and caudate of 9% for 99mTc agents and 19% for 123I agents. Improvements averaged 31% in the putamen and 43% in the caudate when correction for attenuation, scatter and spatially variant collimator response was included in the reconstruction.

  2. Radiation shielding for the Main Injector collimation system

    SciTech Connect

    Rakhno, Igor; /Fermilab

    2008-05-01

    The results of Monte Carlo radiation shielding studies performed with the MARS15 code for the Main Injector collimation system at Fermilab are presented and discussed. MAD-to-MARS Beam Line Builder is used to generate realistic extended curvilinear geometry models.

  3. SNS Collimating System Design — Performance and Integration

    NASA Astrophysics Data System (ADS)

    Simos, N.; Ludewig, H.; Raparia, D.; Catalan-Lasheras, N.; Brodowski, J.; Murdoch, G.

    2003-12-01

    The collimating system in the accumulator ring and transfer lines of the Spallation Neutron Source (SNS) project is responsible for stopping 0.1% of the 2 MW beam of 1.0 GeV protons that are in the beam halo. The collimating structures are a combination of movable beam scrapers and stationary absorbers. Specifically, pairs of charge-exchange foils or scrapers moving in-and-out of the beam in the vertical and horizontal directions help guide the halo protons into respective absorbers which consist of an intricate design of a double wall beam tube, a water-cooled particle bed and radial shielding. Off-momentum protons, with the help of respective charge exchange foils and a dipole magnet, are directed to a momentum dump consisting of a cooled particle bed downstream of a double-walled window separating it from the vacuum space. Addressed in this paper is the thermo-mechanical response and survivability of key components of the collimating system (such as the collimating beam tube in the absorbers, the beam windows and the primary element of the bean scraper structure) in the event of intercept of the full beam under accident conditions. While the potential for the full beam to be intercepted by these components is remote, still special attention will be paid in assessing the amount of full beam (or number of pulses) they can tolerate.

  4. DOUBLE-WALL COLLIMATOR DESIGN OF THE SNS PROJECT.

    SciTech Connect

    SIMOS,N.; LUDEWIG,H.; CATALAN-LASHERAS,N.; CRIVELLO,S.

    2001-06-18

    The collimator absorber array of the Spallation Neutron Source (SNS) project is responsible for stopping the 1.0 GeV protons that are in the halo of the beam. It is estimated that 0.1% of the 2 MW beam will be intercepted by the adopted collimating scheme implemented at various sections of the beam transport and accumulation. This paper summarizes the conceptual design of the collimator absorber as well as the supporting detailed analysis that were performed and guided the design process. Key requirement in the design process is the need for the collimator beam tube to minimize beam impedance while closely following its beta function. Due to lack of available experimental data, the long-term behavior of irradiated materials in an environment where they interface with coolant flow becomes an issue. Uncertainties in the long-term behavior prompted a special double-wall design that will enable not only beam halo interception but also the efficient transfer of deposited energy both under normal and off-normal conditions to the coolant flow. The thermo-mechanical response of the double wall beam tube and of a particle bed surrounding it are discussed in detail in the paper.

  5. The practical Pomeron for high energy proton collimation

    NASA Astrophysics Data System (ADS)

    Appleby, R. B.; Barlow, R. J.; Molson, J. G.; Serluca, M.; Toader, A.

    2016-10-01

    We present a model which describes proton scattering data from ISR to Tevatron energies, and which can be applied to collimation in high energy accelerators, such as the LHC and FCC. Collimators remove beam halo particles, so that they do not impinge on vulnerable regions of the machine, such as the superconducting magnets and the experimental areas. In simulating the effect of the collimator jaws it is crucial to model the scattering of protons at small momentum transfer t, as these protons can subsequently survive several turns of the ring before being lost. At high energies these soft processes are well described by Pomeron exchange models. We study the behaviour of elastic and single-diffractive dissociation cross sections over a wide range of energy, and show that the model can be used as a global description of the wide variety of high energy elastic and diffractive data presently available. In particular it models low mass diffraction dissociation, where a rich resonance structure is present, and thus predicts the differential and integrated cross sections in the kinematical range appropriate to the LHC. We incorporate the physics of this model into the beam tracking code MERLIN and use it to simulate the resulting loss maps of the beam halo lost in the collimators in the LHC.

  6. The collimation of magnetic jets by disc winds

    NASA Astrophysics Data System (ADS)

    Globus, N.; Levinson, A.

    2016-09-01

    The collimation of a Poynting-flux dominated jet by a wind emanating from the surface of an accretion flow is computed using a semi-analytic model. The injection of the disc wind is treated as a boundary condition in the equatorial plane, and its evolution is followed by invoking a prescribed geometry of streamlines. Solutions are obtained for a wide range of disc wind parameters. It is found that jet collimation generally occurs when the total wind power exceeds about 10 percents of the jet power. For moderate wind powers, we find gradual collimation. For strong winds, we find rapid collimation followed by focusing of the jet, after which it remains narrow over many Alfvén crossing times before becoming conical. We estimate that in the later case, the jet's magnetic field may be dissipated by the current-driven kink instability over a distance of a few hundreds gravitational radii. We apply the model to M87 and show that the observed parabolic shape of the radio jet within the Bondi radius can be reproduced provided that the wind injection zone extends to several hundreds gravitational radii, and that its total power is about one-third of the jet power. The radio spectrum can be produced by synchrotron radiation of relativistically hot, thermal electrons in the sheath flow surrounding the inner jet.

  7. Creating a collimated ultrasound beam in highly attenuating fluids.

    PubMed

    Raeymaekers, Bart; Pantea, Cristian; Sinha, Dipen N

    2012-04-01

    We have devised a method, based on a parametric array concept, to create a low-frequency (300-500 kHz) collimated ultrasound beam in fluids highly attenuating to sound. This collimated beam serves as the basis for designing an ultrasound visualization system that can be used in the oil exploration industry for down-hole imaging in drilling fluids. We present the results of two different approaches to generating a collimated beam in three types of highly attenuating drilling mud. In the first approach, the drilling mud itself was used as a nonlinear mixing medium to create a parametric array. However, the short absorption length in mud limits the mixing length and, consequently, the resulting beam is weak and broad. In the second improved approach, the beam generation process was confined to a separate "frequency mixing tube" that contained an acoustically non-linear, low attenuation medium (e.g., water) that allowed establishing a usable parametric array in the mixing tube. A low-frequency collimated beam was thus created prior to its propagation into the drilling fluid. Using the latter technique, the penetration depth of the low frequency ultrasound beam in the drilling fluid was significantly extended. We also present measurements of acoustic nonlinearity in various types of drilling mud. PMID:22204917

  8. Method for Measuring Collimator-Pointing Sensitivity to Temperature Changes

    NASA Technical Reports Server (NTRS)

    Abramovici, Alex; Cox, Timothy E.; Hein, Randall C.; MacDonald, Daniel R.

    2011-01-01

    For a variety of applications, it is important to measure the sensitivity of the pointing of a beam emerging from a collimator, as a function of temperature changes. A straightforward method for carrying out this measurement is based on using interferometry for monitoring the changes in beam pointing, which presents its own problems. The added temperature dependence and complexity issues relating to using an interferometer are addressed by not using an interferometer in the first place. Instead, the collimator is made part of an arrangement that uses a minimum number of low-cost, off-the-shelf materials and by using a quad diode to measure changes in beam pointing. In order to minimize the influence of the test arrangement on the outcome of the measurement, several steps are taken. The collimator assembly is placed on top of a vertical, 1-m-long, fused silica tube. The quad diode is bonded to a fused silica bar, which, in turn, is bonded to the lower end of the fused silica tube. The lower end of the tube rests on a self-aligning support piece, while the upper end of the tube is kept against two rounded setscrew tips, using a soft rubber string. This ensures that very little stress is applied to the tube as the support structure changes dimensions due to thermal expansion. Light is delivered to the collimator through a bare fiber in order to minimize variable bending torque caused by a randomly relaxing, rigid fiber jacket. In order to separate the effect of temperature on the collimator assembly from the effect temperature has on the rest of the setup, multiple measurements are taken with the collimator assembly rotated from measurement to measurement. Laboratory testing, with 1-m spacing between the collimator and the quad diode, has shown that the sensitivity of the arrangement is better than 100 nm rms, over time spans of at least one hour, if the beam path is protected from atmospheric turbulence by a tube. The equivalent sensitivity to detecting changes in

  9. Slow axis collimation lens with variable curvature radius for semiconductor laser bars

    NASA Astrophysics Data System (ADS)

    Xiong, Ling-Ling; Cai, Lei; Zheng, Yan-Fang; Liu, Hui; Zhang, Pu; Nie, Zhi-Qiang; Liu, Xing-Sheng

    2016-03-01

    Based on Snell's law and the constant phase in the front of optical field, a design method of the slow axis collimation lens with variable curvature radius is proposed for semiconductor laser bars. Variable radius of the collimator is designed by the transmission angle, and it is demonstrated that the collimator has good beam collimation ability by material with low refractive index. Resorting to the design thought of finite element method, the surface of the collimator has been divided, and it is feasible to be fabricated. This method is applied as an example in collimation of a 976 nm semiconductor laser bar. 6 mrad divergence angle of collimated beam at slow axis is realized by the designed collimation lens with refraction index of 1.51.

  10. Observation of strong leakage reduction in crystal assisted collimation of the SPS beam

    NASA Astrophysics Data System (ADS)

    Scandale, W.; Arduini, G.; Butcher, M.; Cerutti, F.; Garattini, M.; Gilardoni, S.; Lechner, A.; Losito, R.; Masi, A.; Mereghetti, A.; Metral, E.; Mirarchi, D.; Montesano, S.; Redaelli, S.; Rossi, R.; Schoofs, P.; Smirnov, G.; Bagli, E.; Bandiera, L.; Baricordi, S.; Dalpiaz, P.; Germogli, G.; Guidi, V.; Mazzolari, A.; Vincenzi, D.; Claps, G.; Dabagov, S.; Hampai, D.; Murtas, F.; Cavoto, G.; Iacoangeli, F.; Ludovici, L.; Santacesaria, R.; Valente, P.; Galluccio, F.; Afonin, A. G.; Chesnokov, Yu. A.; Durum, A. A.; Maisheev, V. A.; Sandomirskiy, Yu. E.; Yanovich, A. A.; Kovalenko, A. D.; Taratin, A. M.; Gavrikov, Yu. A.; Ivanov, Yu. M.; Lapina, L. P.; Fulcher, J.; Hall, G.; Pesaresi, M.; Raymond, M.

    2015-09-01

    In ideal two-stage collimation systems, the secondary collimator-absorber should have its length sufficient to exclude practically the exit of halo particles with large impact parameters. In the UA9 experiments on the crystal assisted collimation of the SPS beam a 60 cm long tungsten bar is used as a secondary collimator-absorber which is insufficient for the full absorption of the halo protons. Multi-turn simulation studies of the collimation allowed to select the position for the beam loss monitor downstream the collimation area where the contribution of particles deflected by the crystal in channeling regime but emerging from the secondary collimator-absorber is considerably reduced. This allowed observation of a strong leakage reduction of halo protons from the SPS beam collimation area, thereby approaching the case with an ideal absorber.

  11. The manipulation of self-collimated beam in phononic crystals composed of orientated rectangular inclusions

    NASA Astrophysics Data System (ADS)

    Tsai, Chia-Nien; Chen, Lien-Wen

    2016-07-01

    Self-collimation is wave propagation in straight path without diffraction. The performance is evaluated by bandwidth, angular collimating range and straightness of equi-frequency contours. The present study aims to manipulate the self-collimated beam in square-array phononic crystals by means of orientated rectangular inclusions. Finite element simulations are performed to investigate the effects of the aspect ratio and orientation angle of rectangular inclusions on the self-collimated beam. The simulation results show that the proposed design successfully achieves all-angle self-collimation phenomenon. In addition, it also shows that the propagation direction of a self-collimated beam can be effectively manipulated by varying the orientation angle of inclusions. Numerical simulation result of the S-shaped bend demonstrates that acoustic collimated beam can be steered with negligible diffraction. Overall, the proposed design has significant potential for the realization of applications such as collimators, acoustic waveguides and other phononic crystals-based systems.

  12. Controlling the collimation and rotation of hydromagnetic disc winds

    NASA Astrophysics Data System (ADS)

    Pudritz, Ralph E.; Rogers, Conrad S.; Ouyed, Rachid

    2006-02-01

    We present a comprehensive set of axisymmetric, time-dependent simulations of jets from Keplerian discs whose mass loading as a function of disc radius is systematically changed. For a reasonable model for the density structure and injection speed of the underlying accretion disc, mass loading is determined by the radial structure of the magnetic field of the disc. We vary this structure by using four different magnetic field configurations, ranging from the Ouyed-Pudritz `potential' configuration, to the increasingly more steeply falling Blandford-Payne and Pelletier-Pudritz models, and ending with a quite steeply raked configuration that bears similarities to the Shu X-wind model. We find that the radial distribution of the mass load has a profound effect on the rotational profile of the underlying jet as well as the degree of collimation of its outflow velocity and magnetic field lines. These four models have systematic differences in the power-law rotation profiles of jet material far from the source: vφ(r)~ra ranging over -0.46 >=a>=-0.76. We show analytically and confirm by our simulations that the collimation of a jet depends on its radial current distribution, which in turn is prescribed by the mass load. Models with steeply descending mass loads have strong toroidal fields, and these collimate to cylinders (this includes the Ouyed-Pudritz and Blandford-Payne outflows). On the other hand, the more gradually descending mass load profiles (the PP92 and monopolar distributions) have weaker toroidal fields and these result in wide-angle outflows with parabolic collimation. We also present detailed structural information about jets such as their radial profiles of jet density, toroidal magnetic field and poloidal jet speed, as well as an analysis of the bulk energetics of our different simulations. Our results are in excellent agreement with the predictions of asymptotic collimation for axisymmetric, stationary jets.

  13. Photon absorptiometry

    SciTech Connect

    Velchik, M.G.

    1987-01-01

    Recently, there has been a renewed interest in the detection and treatment of osteoporosis. This paper is a review of the merits and limitations of the various noninvasive modalities currently available for the measurement of bone mineral density with special emphasis placed upon the nuclear medicine techniques of single-photon and dual-photon absorptiometry. The clinicians should come away with an understanding of the relative advantages and disadvantages of photon absorptiometry and its optimal clinical application. 49 references.

  14. Poster — Thur Eve — 64: Preliminary investigation of arc configurations for optimal sparing of normal tissue in hypofractionated stereotactic radiotherapy (HF-SRT) of multiple brain metastases using a 5mm interdigitating micro-multileaf collimator

    SciTech Connect

    Leavens, C; Wronski, M; Lee, YK; Ruschin, M; Soliman, H; Sahgal, A

    2014-08-15

    Purpose: To evaluate normal tissue sparing in intra-cranial HF-SRT, comparing various arc configurations with the Synergy Beam Modulator (SynBM) and Agility linacs, the latter incorporating leaf interdigitation and backup jaws. Methods: Five patients with multiple brain metastases (BMs), (5 BMs (n=2), 3 BMs (n=3)) treated with HF-SRT using 25 Gy (n=2) or 30 Gy (n=3) in 5 fractions, were investigated. Clinical treatment plans used the SynBM. Each patient was retrospectively re-planned on Agility, employing three planning strategies: (A) one isocenter and dedicated arc for each BM; (B) a single isocenter, centrally placed with respect to BMs; (C) the isocenter and arc configuration used in the SynBM plan, where closely spaced (<5cm) BMs used a dedicated isocenter and arcs. Agility plans were normalized for PTV coverage and heterogeneity. Results and Conclusion: Strategy A obtained the greatest improvements over the SynBM plan, where the maximum OAR dose, and mean dose to normal brain (averaged for all patients) were reduced by 55cGy and 25cGy, respectively. Strategy B was limited by having a single isocenter, hence less jaw shielding and increased MLC leakage. The maximum OAR dose was reduced by 13cGy, however mean dose to normal brain increased by 84cGy. Strategy C reduced the maximum OAR dose and mean dose to normal brain by 32cGy and 9cGy, respectively. The results from this study indicate that, for intra-cranial HF-SRT of multiple BMs, Agility plans are equal or better than SynBM plans. Further planning is needed to investigate dose sparing using Strategy A and the SynBM.

  15. Comparison of Techniques to Reduce Bremsstrahlung Background Radiation from Monoenergetic Photon Beams

    SciTech Connect

    Johnson, M; McNabb, D

    2006-06-29

    An important applied technology is a tunable mono-energetic photon source [1]. These sources are made of relativistic electron accelerators coupled to low-energy lasers, which produce high-energy, mono-energetic-rays. One challenge associated with systems such as this is a continuum of bremsstrahlung background created when an electron beam passes through an aperture of some sort and the electron bunch or its halo impinges on the aperture pictured in figure 1. For instance, in the current T-REX [1] design for the interaction point between the laser- and electron-beam, the electron-beam passes through the center of a mirror used to reflect the laser. There is a potential with this design that bremsstrahlung radiation may be produced at the edges of the mirror openings and contaminate the mono-energetic photon beam. Certain applications [2] may be sensitive to this contamination. To reduce the bremsstrahlung contaminate a collimator (thickness {approx}24in. (calculated from XCOM database [3]) to attenuate by a factor of 10{sup -3} the 112MeV photons expected in the T-REX demonstration [1]) is situated between the aperture and target. To maximize the brightness of the photon-beam, the collimator opening must be no less than the size of the photon-beam spot size expected to be about 1mm. This fixes the collimator opening. a priori the aperture size must be greater than the collimator opening and is a function distance between the aperture and collimator. In this paper we focus on two approaches to estimate the aperture size, given a collimator and a target whose sizes and distances from the aperture are given. In the next section we will discuss these approaches.

  16. The photon

    NASA Astrophysics Data System (ADS)

    Collins, Russell L.

    2009-10-01

    There are no TEM waves, only photons. Lets build a photon, using a radio antenna. A short antenna (2L<< λ) simplifies the calculation, letting B fall off everywhere as 1/r^2. The Biot-Savart law finds B = (μ0/4π)(LI0/r^2)θφt. The magnetic flux thru a semi-circle of radius λ/2 is set equal to the flux quantum h/e, determining the needed source strength, LI0. From this, one can integrate the magnetic energy density over a sphere of radius λ/2 and finds it to be 1.0121 hc/λ. Pretty close. A B field collapses when the current ceases, but the photon evades this by creating a ɛ0E / t displacement current at center that fully supports the toroidal B assembly as it moves at c. This E=vxB arises because the photon moves at c. Stopped, a photon decays. At every point along the photon's path, an observer will note a transient oscillation of an E field. This sources the EM ``guiding wave'', carrying little or no energy and expanding at c. At the head of the photon, all these spherical guiding waves gather ``in-phase'' as a planar wavefront. This model speaks to all the many things we know about light. The photon is tiny, but its guiding wave is huge.

  17. Collimation system for beam loss localization with slip stacking injection in the Fermilab Main Injector

    SciTech Connect

    Brown, Bruce C.; /Fermilab

    2008-09-01

    Slip stacking injection for high intensity operation of the Fermilab Main Injector produces a small fraction of beam which is not captured in buckets and accelerated. A collimation system has been implemented with a thin primary collimator to define the momentum aperture at which this beam is lost and four massive secondary collimators to capture the scattered beam. The secondary collimators define tight apertures and thereby capture a fraction of other lost beam. The system was installed in 2007 with commissioning continuing in 2008. The collimation system will be described including simulation, design, installation, and commissioning. Successful operation and operational limitations will be described.

  18. Innovative light collimator with afocal lens and total internal reflection lens for daylighting system.

    PubMed

    Chen, Bo-Jian; Chen, Yin-Ti; Ullah, Irfan; Chou, Chun-Han; Chan, Kai-Cyuan; Lai, Yi-Lung; Lin, Chia-Ming; Chang, Cheng-Ming; Whang, Allen Jong-Woei

    2015-10-01

    This research presents a novel design of the collimator, which uses total internal reflection (TIR), convex, and concave lenses for the natural light illumination system (NLIS). The concept of the NLIS is to illuminate building interiors with natural light, which saves energy consumption. The TIR lens is used to collimate the light, and convex and concave lenses are used to converge the light to the required area. The results have shown that the efficiency in terms of achieving collimated light using the proposed collimator at the output of the light collector is better than that of a previous system without a collimator. PMID:26479648

  19. A radial collimator for a time-of-flight neutron spectrometer

    SciTech Connect

    Stone, M. B.; Abernathy, D. L.; Niedziela, J. L.; Loguillo, M. J.; Overbay, M. A.

    2014-08-15

    We have engineered and installed a radial collimator for use in the scattered beam of a neutron time-of-flight spectrometer at a spallation neutron source. The radial collimator may be used with both thermal and epithermal neutrons, reducing the detected scattering intensity due to material outside of the sample region substantially. The collimator is located inside of the sample chamber of the instrument, which routinely cycles between atmospheric conditions and cryogenic vacuum. The oscillation and support mechanism of the collimator allow it to be removed from use without breaking vacuum. We describe here the design and characterization of this radial collimator.

  20. Collimation and splitting of valley electron diffraction in graphene

    NASA Astrophysics Data System (ADS)

    Yang, Mou; Bai, Yan-Kui; Zhang, Wen-Lian; Wang, Rui-Qiang

    2016-08-01

    We reported the collimation and splitting effects of the diffraction of valley electrons in graphene. When the incident energy increases from the neutral point, the diffraction tends to be collimated for one valley and split for the other valley. The difference in the diffraction between valleys results in valley-dependent transport. We investigated the left-right conductance of a four-terminal graphene device. The conductance ratio between the two valleys was derived to be 1 -(8 /3 )E , where E is the incident energy in units of the atom-atom hopping. The ratio is independent of the device dimensions and reflects the intrinsic properties of the electronic structure of graphene.

  1. Ocular vergence measurement in projected and collimated simulator displays.

    PubMed

    Morahan, P; Meehan, J W; Patterson, J; Hughes, P K

    1998-09-01

    The purpose of this study was to investigate electrooculography (EOG) as a measurement of ocular vergence in both collimated and projected simulator environments. The task required participants to shift their gaze between a central fixation point and a target appearing at one of three eccentricities. EOG was effective in recording ocular vergence. The EOG results were similar between collimated and projected displays, except for differences in vergence changes during lateral movement of the eyes, and ocular excursions downward elicited a greater EOG response than the reverse upward movement. The computer-based technique of recording vergence was found to produce measurable traces from a majority of participants. The technique has potential for further development as a tool for measuring ocular vergence in virtual environments where methods that require the wearing of head-mounted apparatus to track ocular structures (e.g., the pupil), which cannot be worn at the same time as a flight or flight-simulator helmet, are unsuitable.

  2. Simulator for beam-based LHC collimator alignment

    NASA Astrophysics Data System (ADS)

    Valentino, Gianluca; Aßmann, Ralph; Redaelli, Stefano; Sammut, Nicholas

    2014-02-01

    In the CERN Large Hadron Collider, collimators need to be set up to form a multistage hierarchy to ensure efficient multiturn cleaning of halo particles. Automatic algorithms were introduced during the first run to reduce the beam time required for beam-based setup, improve the alignment accuracy, and reduce the risk of human errors. Simulating the alignment procedure would allow for off-line tests of alignment policies and algorithms. A simulator was developed based on a diffusion beam model to generate the characteristic beam loss signal spike and decay produced when a collimator jaw touches the beam, which is observed in a beam loss monitor (BLM). Empirical models derived from the available measurement data are used to simulate the steady-state beam loss and crosstalk between multiple BLMs. The simulator design is presented, together with simulation results and comparison to measurement data.

  3. Outflow Propagation in Collapsars: Collimated Jets And Expanding Outflows

    SciTech Connect

    Mizuta, A.; Yamasaki, T.; Nagataki, S.; Mineshige, S.; /Kyoto U., Yukawa Inst., Kyoto /KIPAC, Menlo Park

    2007-06-08

    We investigate the outflow propagation in the collapsar in the context of gamma-ray bursts (GRBs) with 2D relativistic hydrodynamic simulations. We vary the specific internal energy and bulk Lorentz factor of the injected outflow from non-relativistic regime to relativistic one, fixing the power of the outflow to be 10{sup 51}erg s{sup -1}. We observed the collimated outflow, when the Lorentz factor of the injected outflow is roughly greater than 2. To the contrary, when the velocity of the injected outflow is slower, the expanding outflow is observed. The transition from collimated jet to expanding outflow continuously occurs by decreasing the injected velocity. Different features of the dynamics of the outflows would cause the difference between the GRBs and similar phenomena, such as, X-ray flashes.

  4. Experimental demonstration of an inertial collimation mechanism in nested outflows.

    PubMed

    Yurchak, R; Ravasio, A; Pelka, A; Pikuz, S; Falize, E; Vinci, T; Koenig, M; Loupias, B; Benuzzi-Mounaix, A; Fatenejad, M; Tzeferacos, P; Lamb, D Q; Blackman, E G

    2014-04-18

    Interaction between a central outflow and a surrounding wind is common in astrophysical sources powered by accretion. Understanding how the interaction might help to collimate the inner central outflow is of interest for assessing astrophysical jet formation paradigms. In this context, we studied the interaction between two nested supersonic plasma flows generated by focusing a long-pulse high-energy laser beam onto a solid target. A nested geometry was created by shaping the energy distribution at the focal spot with a dedicated phase plate. Optical and x-ray diagnostics were used to study the interacting flows. Experimental results and numerical hydrodynamic simulations indeed show the formation of strongly collimated jets. Our work experimentally confirms the "shock-focused inertial confinement" mechanism proposed in previous theoretical astrophysics investigations. PMID:24785042

  5. 'Crystal Collimator' Measurement of CESR particle-beam Source Size

    SciTech Connect

    Finkelstein, K.D.; Bazarov, Ivan; White, Jeffrey; Revesz, Peter

    2004-05-12

    We have measured electron and positron beam source size at CHESS when the Cornell Electron Storage Ring (CESR) is run dedicated for the production of synchrotron radiation. Horizontal source size at several beamlines is expected to shrink by a factor of two but synchrotron (visible) light measurements only provide the vertical size. Therefore a 'crystal collimator' using two Bragg reflection in dispersive (+,+) orientation has been built to image the horizontal (vertical) source by passing x-rays parallel to within 5 microradians to an imaging screen and camera. With the 'crystal collimator' we observe rms sizes of 1.2 mm horizontal by 0.28 mm vertical, in good agreement with the 1.27 mm size calculated from lattice functions, and 0.26 mm observed using a synchrotron light interferometer.

  6. First results on the SPS beam collimation with bent crystals

    SciTech Connect

    Scandale, W.; Robert-Demolaize, G.; Arduini, G.; Assmann, R.; Bracco, C.; Gilardoni, S.; Ippolito, V.; Laface, E.; Losito, R,; Masi, A,; Metral, E,; Previtali, V.; Redaelli, S.; Silari, M.; Tlustos, L.; Bagli, E.; Baricordi, S.; Dalpiaz, P.; Guidi, V.; Mazzolari, A.; Vincenzi, D.; Della Mea, G.; Lombardi, A.; De Salvador, D.; Vallazza, E.; Bolognini, D.; Hasan, S.; Lietti, D.; Mascagna, V.; Mattera, A.; Prest, M.; Cavoto, G.; Ludovici, L.; Mirarchi, D.; Santacesaria, R.; Valente, P.; Murtas, F.; Afonin, A.G.; Chesnokov, Yu.A.; Maisheev, V.A.; Yazynin, I.A.; Kovalenko, A.D.; Taratin, A.M.; Denisov, A.S.; Gavrikov, Yu.A.; Ivanov, Yu.M.; Lapina, L.P.; Malyarenko, L.G.; Skorobogatov, V.V.; Suvorov, V.M.; Vavilov, S.A.; Mokhov, N.; Still, D.; Markiewicz, T.; Oriunno, M.

    2010-08-23

    Experiments were performed at the CERN SPS with 120 GeV/c stored proton beams to assess the possibility of beam halo collimation assisted by bent crystals. A bent crystal was used to deflect horizontally by an angle of about 170 {micro}rad the beam halo protons in channeling states directing them into a 60 cm long tungsten absorber. The halo loss rate due to nuclear inelastic interactions of protons in the aligned crystal was up to five times smaller than for its amorphous orientation. Channeled fractions, (75 {+-} 4)% and (85 {+-} 5)% for the two tested silicon crystals, were measured by intercepting the deflected beam with another collimator located between the crystals and the absorber. The pixel detector (MEDIPIX) installed in a Roman pot inside the beam pipe was used to obtain visual images of the deflected beam.

  7. Experimental Demonstration of an Inertial Collimation Mechanism in Nested Outflows

    NASA Astrophysics Data System (ADS)

    Yurchak, R.; Ravasio, A.; Pelka, A.; Pikuz, S.; Falize, E.; Vinci, T.; Koenig, M.; Loupias, B.; Benuzzi-Mounaix, A.; Fatenejad, M.; Tzeferacos, P.; Lamb, D. Q.; Blackman, E. G.

    2014-04-01

    Interaction between a central outflow and a surrounding wind is common in astrophysical sources powered by accretion. Understanding how the interaction might help to collimate the inner central outflow is of interest for assessing astrophysical jet formation paradigms. In this context, we studied the interaction between two nested supersonic plasma flows generated by focusing a long-pulse high-energy laser beam onto a solid target. A nested geometry was created by shaping the energy distribution at the focal spot with a dedicated phase plate. Optical and x-ray diagnostics were used to study the interacting flows. Experimental results and numerical hydrodynamic simulations indeed show the formation of strongly collimated jets. Our work experimentally confirms the "shock-focused inertial confinement" mechanism proposed in previous theoretical astrophysics investigations.

  8. Dense Plasma Focus as Collimated Source of D-D Fusion Neutron Beams for Irradiation Experiences and Study of Emitted Radiations

    SciTech Connect

    Milanese, M.; Niedbalski, J.; Moroso, R.; Guichon, S.; Supan, J.

    2008-04-07

    A 'table-top' 2 kJ, 250 kA plasma focus, the PACO (Plasma AutoConfinado), designed by the Dense Plasma Group of IFAS is used in its optimum regime for neutron yield for obtaining collimated pulsed neutron beams (100 ns). A simple and low-cost shielding arrangement was developed in order to fully eliminate the 2.45 MeV neutrons generated in the PACO device (10{sup 8} per shot at 31 kV, 1-2 mbar). Conventional neutron diagnostics: scintillator-photomultiplier (S-PMT), silver activation counters (SAC), etc., are used to determine the minimum width of the shielding walls. Emission of very hard electromagnetic pulses is also studied. Collimation using lead and copper plates is made to determine the localization of the very hard X-ray source. The maximum energy of the continuum photon distribution is estimated in 0,6 MeV using a system of filters.

  9. Changes to the Transfer Line Collimation System for the High-Luminosity LHC Beams

    SciTech Connect

    Kain, V.; Aberle, O.; Bracco, C.; Fraser, M.; Galleazzi, F.; Gianfelice-Wendt, E.; Kosmicki, A.; Maciariello, F.; Meddahi, M.; Nuiry, F. X.; Steele, G.; Velotti, F.

    2015-06-01

    The current LHC transfer line collimation system will not be able to provide enough protection for the high brightness beams in the high-luminosity LHC era. The new collimation system will have to attenuate more and be more robust than its predecessor. The active jaw length of the new transfer line collimators will therefore be 2.1 m instead of currently 1.2 m. The transfer line optics will have to be adjusted for the new collimator locations and larger beta functions at the collimators for absorber robustness reasons. In this paper the new design of the transfer line collimation system will be presented with its implications on transfer line optics and powering, maintainability, protection of transfer line magnets in case of beam loss on a collimator and protection of the LHC aperture.

  10. Device for detachably securing a collimator to a radiation detector

    SciTech Connect

    Hanz, G.J.; Jung, G.; Pflaum, M.

    1986-12-16

    A device is described for detachably securing a collimator to a radiation detector, comprising: (a) a first annular groove means secured to the radiation detector; (b) a second annular groove means secured to the collimator; (c) a split ring having a first and second ring ends, the ring being received in the first annular groove means; and (d) a ring diameter control system, including (d1) a first lever system having two ends; (d2) a second lever system having two ends; and (d3) a rotating hub being rotatably secured to the detector head; wherein the first lever system is rotatably mounted with one end linked to the first ring end and with the other end linked to the rotating hub. The second lever system is rotatably mounted with one end linked to the second ring end and with the other end linked to the rotating hub, such that rotation of the rotating hub moves the first and second lever systems in opposite directions thereby moving the first and second ring ends between a first position, in which the split ring is positioned only in the first annular groove means, and a second position, in which the split ring is located in both the first annular groove means and the second annular groove means, thus attaching the collimator to the radiation detector.

  11. SYNCHROTRON RADIATION OF SELF-COLLIMATING RELATIVISTIC MAGNETOHYDRODYNAMIC JETS

    SciTech Connect

    Porth, Oliver; Fendt, Christian; Vaidya, Bhargav; Meliani, Zakaria E-mail: fendt@mpia.de

    2011-08-10

    The goal of this paper is to derive signatures of synchrotron radiation from state-of-the-art simulation models of collimating relativistic magnetohydrodynamic (MHD) jets featuring a large-scale helical magnetic field. We perform axisymmetric special relativistic MHD simulations of the jet acceleration region using the PLUTO code. The computational domain extends from the slow-magnetosonic launching surface of the disk up to 6000{sup 2} Schwarzschild radii allowing jets to reach highly relativistic Lorentz factors. The Poynting-dominated disk wind develops into a jet with Lorentz factors of {Gamma} {approx_equal} 8 and is collimated to 1{sup 0}. In addition to the disk jet, we evolve a thermally driven spine jet emanating from a hypothetical black hole corona. Solving the linearly polarized synchrotron radiation transport within the jet, we derive very long baseline interferometry radio and (sub-) millimeter diagnostics such as core shift, polarization structure, intensity maps, spectra, and Faraday rotation measure (RM) directly from the Stokes parameters. We also investigate depolarization and the detectability of a {lambda}{sup 2}-law RM depending on beam resolution and observing frequency. We find non-monotonic intrinsic RM profiles that could be detected at a resolution of 100 Schwarzschild radii. In our collimating jet geometry, the strict bimodality in the polarization direction (as predicted by Pariev et al.) can be circumvented. Due to relativistic aberration, asymmetries in the polarization vectors across the jet can hint at the spin direction of the central engine.

  12. First Sub-arcsecond Collimation of Monochromatic Neutrons

    NASA Astrophysics Data System (ADS)

    Wagh, Apoorva G.; Abbas, Sohrab; Treimer, Wolfgang

    2010-11-01

    We have achieved the tightest collimation to date of a monochromatic neutron beam by diffracting neutrons from a Bragg prism, viz. a single crystal prism operating in the vicinity of Bragg incidence. An optimised silicon {111} Bragg prism has collimated 5.26Å neutrons down to 0.58 arcsecond. In conjunction with a similarly optimised Bragg prism analyser of opposite asymmetry, this ultra-parallel beam yielded a 0.62 arcsecond wide rocking curve. This beam has produced the first SUSANS spectrum in Q ~ 10-6 Å-1 range with a hydroxyapatite casein protein sample and demonstrated the instrument capability of characterising agglomerates upto 150 μm in size. The super-collimation has also enabled recording of the first neutron diffraction pattern from a macroscopic grating of 200 μm period. An analysis of this pattern yielded the beam transverse coherence length of 175 μm (FWHM), the greatest achieved to date for Å wavelength neutrons.

  13. LOR-interleaving image reconstruction for PET imaging with fractional-crystal collimation

    NASA Astrophysics Data System (ADS)

    Li, Yusheng; Matej, Samuel; Karp, Joel S.; Metzler, Scott D.

    2015-01-01

    Positron emission tomography (PET) has become an important modality in medical and molecular imaging. However, in most PET applications, the resolution is still mainly limited by the physical crystal sizes or the detector’s intrinsic spatial resolution. To achieve images with better spatial resolution in a central region of interest (ROI), we have previously proposed using collimation in PET scanners. The collimator is designed to partially mask detector crystals to detect lines of response (LORs) within fractional crystals. A sequence of collimator-encoded LORs is measured with different collimation configurations. This novel collimated scanner geometry makes the reconstruction problem challenging, as both detector and collimator effects need to be modeled to reconstruct high-resolution images from collimated LORs. In this paper, we present a LOR-interleaving (LORI) algorithm, which incorporates these effects and has the advantage of reusing existing reconstruction software, to reconstruct high-resolution images for PET with fractional-crystal collimation. We also develop a 3D ray-tracing model incorporating both the collimator and crystal penetration for simulations and reconstructions of the collimated PET. By registering the collimator-encoded LORs with the collimator configurations, high-resolution LORs are restored based on the modeled transfer matrices using the non-negative least-squares method and EM algorithm. The resolution-enhanced images are then reconstructed from the high-resolution LORs using the MLEM or OSEM algorithm. For validation, we applied the LORI method to a small-animal PET scanner, A-PET, with a specially designed collimator. We demonstrate through simulated reconstructions with a hot-rod phantom and MOBY phantom that the LORI reconstructions can substantially improve spatial resolution and quantification compared to the uncollimated reconstructions. The LORI algorithm is crucial to improve overall image quality of collimated PET, which

  14. LOR-interleaving image reconstruction for PET imaging with fractional-crystal collimation

    PubMed Central

    Li, Yusheng; Matej, Samuel; Karp, Joel S.; Metzler, Scott D.

    2015-01-01

    Positron emission tomography (PET) has become an important modality in medical and molecular imaging. However, in most PET applications, the resolution is still mainly limited by the physical crystal sizes or the detector’s intrinsic spatial resolution. To achieve images with better spatial resolution in a central region of interest (ROI), we have previously proposed using collimation in PET scanner. The collimator is designed to partially mask detector crystals to detect lines of response (LORs) within fractional crystals. A sequence of collimator-encoded LORs is measured with different collimation configurations. This novel collimated scanner geometry makes the reconstruction problem challenging, as both detector and collimator effects need to be modeled to reconstruct high-resolution images from collimated LORs. In this paper, we present an LOR-interleaving (LORI) algorithm, which incorporates these effects and has the advantage of reusing existing reconstruction software, to reconstruct high-resolution images for PET with fractional-crystal collimation. We also develop a 3-D ray-tracing model incorporating both the collimator and crystal penetration for simulations and reconstructions of the collimated PET. By registering the collimator-encoded LORs with the collimator configurations, high-resolution LORs are restored based on the modeled transfer matrices using the nonnegative least-squares method and EM algorithm. The resolution-enhanced images are then reconstructed from the high-resolution LORs using the MLEM or OSEM algorithm. For validation, we applied the LORI method to a small-animal PET scanner, A-PET, with a specially designed collimator. We demonstrate through simulated reconstructions with a hot-rod phantom and MOBY phantom that the LORI reconstructions can substantially improve spatial resolution and quantification compared to the uncollimated reconstructions. The LORI algorithm is crucial to improve overall image quality of collimated PET, which

  15. The effect of gamma ray penetration on angle-dependent sensitivity for pinhole collimation in nuclear medicine

    SciTech Connect

    Smith, M.F.; Jaszczak, R.J. |

    1997-11-01

    The sensitivity of a pinhole collimator for gamma ray imaging in nuclear medicine is dependent on the angle of incidence of the gamma rays. The effect of penetration near the pinhole aperture on angle-dependent sensitivity was investigated using experimental measurements and numerical modeling. Projection data measurements were acquired with Tc-99m and I-131 point sources using tungsten pinhole inserts with 1.0 to 4.0 mm diameter apertures. Curves of the form sin{sup x}{theta}, where {theta} is the angle of the incident ray with the surface of the detector crystal, were fit to sensitivity measurements from the projection data. Experimentally measured x values were between 3.3 and 4.1 for Tc-99m and between 5.1 and 7.2 for I-131. Penetration near the pinhole aperture was modeled using (1) an expression for effective pinhole diameter that is a generalization of Anger{close_quote}s formula for normally incident photons and (2) a photon transport simulation code. Experimentally measured sensitivity exponents x from new and previously reported experimental observations were modeled within 15{percent} by the numerical simulations. For modeling using the generalized expression for effective diameter the average error was 1.4{percent} and the standard deviation was 7.7{percent}. For the photon transport simulation code the average error was 1.5{percent} and the standard deviation also was 7.7{percent}. The effect of pinhole aperture design parameters on angle-dependent sensitivity for high resolution pinhole apertures was modeled using a photon transport simulation code. The sensitivity exponents x were greater for 364 keV photons than for 140 keV photons and were greater for small aperture diameters, small acceptance angles, and large aperture channel heights. (Abstract Truncated)

  16. Photon generator

    DOEpatents

    Srinivasan-Rao, Triveni

    2002-01-01

    A photon generator includes an electron gun for emitting an electron beam, a laser for emitting a laser beam, and an interaction ring wherein the laser beam repetitively collides with the electron beam for emitting a high energy photon beam therefrom in the exemplary form of x-rays. The interaction ring is a closed loop, sized and configured for circulating the electron beam with a period substantially equal to the period of the laser beam pulses for effecting repetitive collisions.

  17. Collimating Montel mirror as part of a multi-crystal analyzer system for resonant inelastic X-ray scattering.

    PubMed

    Kim, Jungho; Shi, Xianbo; Casa, Diego; Qian, Jun; Huang, XianRong; Gog, Thomas

    2016-07-01

    Advances in resonant inelastic X-ray scattering (RIXS) have come in lockstep with improvements in energy resolution. Currently, the best energy resolution at the Ir L3-edge stands at ∼25 meV, which is achieved using a diced Si(844) spherical crystal analyzer. However, spherical analyzers are limited by their intrinsic reflection width. A novel analyzer system using multiple flat crystals provides a promising way to overcome this limitation. For the present design, an energy resolution at or below 10 meV was selected. Recognizing that the angular acceptance of flat crystals is severely limited, a collimating element is essential to achieve the necessary solid-angle acceptance. For this purpose, a laterally graded, parabolic, multilayer Montel mirror was designed for use at the Ir L3-absorption edge. It provides an acceptance larger than 10 mrad, collimating the reflected X-ray beam to smaller than 100 µrad, in both vertical and horizontal directions. The performance of this mirror was studied at beamline 27-ID at the Advanced Photon Source. X-rays from a diamond (111) monochromator illuminated a scattering source of diameter 5 µm, generating an incident beam on the mirror with a well determined divergence of 40 mrad. A flat Si(111) crystal after the mirror served as the divergence analyzer. From X-ray measurements, ray-tracing simulations and optical metrology results, it was established that the Montel mirror satisfied the specifications of angular acceptance and collimation quality necessary for a high-resolution RIXS multi-crystal analyzer system. PMID:27359136

  18. Collimating Montel mirror as part of a multi-crystal analyzer system for resonant inelastic X-ray scattering.

    PubMed

    Kim, Jungho; Shi, Xianbo; Casa, Diego; Qian, Jun; Huang, XianRong; Gog, Thomas

    2016-07-01

    Advances in resonant inelastic X-ray scattering (RIXS) have come in lockstep with improvements in energy resolution. Currently, the best energy resolution at the Ir L3-edge stands at ∼25 meV, which is achieved using a diced Si(844) spherical crystal analyzer. However, spherical analyzers are limited by their intrinsic reflection width. A novel analyzer system using multiple flat crystals provides a promising way to overcome this limitation. For the present design, an energy resolution at or below 10 meV was selected. Recognizing that the angular acceptance of flat crystals is severely limited, a collimating element is essential to achieve the necessary solid-angle acceptance. For this purpose, a laterally graded, parabolic, multilayer Montel mirror was designed for use at the Ir L3-absorption edge. It provides an acceptance larger than 10 mrad, collimating the reflected X-ray beam to smaller than 100 µrad, in both vertical and horizontal directions. The performance of this mirror was studied at beamline 27-ID at the Advanced Photon Source. X-rays from a diamond (111) monochromator illuminated a scattering source of diameter 5 µm, generating an incident beam on the mirror with a well determined divergence of 40 mrad. A flat Si(111) crystal after the mirror served as the divergence analyzer. From X-ray measurements, ray-tracing simulations and optical metrology results, it was established that the Montel mirror satisfied the specifications of angular acceptance and collimation quality necessary for a high-resolution RIXS multi-crystal analyzer system.

  19. Photonic lanterns

    NASA Astrophysics Data System (ADS)

    Leon-Saval, Sergio G.; Argyros, Alexander; Bland-Hawthorn, Joss

    2013-12-01

    Multimode optical fibers have been primarily (and almost solely) used as "light pipes" in short distance telecommunications and in remote and astronomical spectroscopy. The modal properties of the multimode waveguides are rarely exploited and mostly discussed in the context of guiding light. Until recently, most photonic applications in the applied sciences have arisen from developments in telecommunications. However, the photonic lantern is one of several devices that arose to solve problems in astrophotonics and space photonics. Interestingly, these devices are now being explored for use in telecommunications and are likely to find commercial use in the next few years, particularly in the development of compact spectrographs. Photonic lanterns allow for a low-loss transformation of a multimode waveguide into a discrete number of single-mode waveguides and vice versa, thus enabling the use of single-mode photonic technologies in multimode systems. In this review, we will discuss the theory and function of the photonic lantern, along with several different variants of the technology. We will also discuss some of its applications in more detail. Furthermore, we foreshadow future applications of this technology to the field of nanophotonics.

  20. Diamond-anvil high-pressure cell with improved x-ray collimation system

    DOEpatents

    Schiferl, D.; Olinger, B.W.; Livingston, R.W.

    1984-03-30

    An adjustable x-ray collimation system for a diamond-anvil high-pressure cell of the type including a cooperable piston and cylinder and a pair of opposing diamonds located between the head of the piston and the head of the cylinder. The x-ray collimation system includes a tubular insert which contains an x-ray collimator. The insert is engageable in the bore of the piston. The collimator is mounted within the insert by means of an elastomeric o-ring at the end closest the opposed diamonds, and by means of a set of adjustable set screws at the opposite end. By adjustment of the set screws the collimator can be pivoted about the o-ring and brought into alignment with the opposed diamonds and the sample contained therein. In the preferred embodiment there is further provided a set of plugs which are insertable in the bore of the collimator. The plugs have bores of different diameters. By successively inserting plugs of progressively smaller bore diameters and adjusting the alignment of the collimator with each plug, the collimator can be quickly brought into accurate alignment with the diamonds. The collimation system allows alignment of the collimator either before or after the cell has been loaded and pressurized.

  1. Diamond-anvil high-pressure cell with improved X-ray collimation system

    DOEpatents

    Schiferl, David; Olinger, Barton W.; Livingston, Robert W.

    1986-01-01

    An adjustable X-ray collimation system for a diamond-anvil high-pressure cell of the type including a cooperable piston and cylinder and a pair of opposing diamonds located between the head of the piston and the head of the cylinder. The X-ray collimation system includes a tubular insert which contains an X-ray collimator. The insert is engageable in the bore of the piston. The collimator is mounted within the insert by means of an elastomeric O-ring at the end closest the opposed diamonds, and by means of a set of adjustable set screws at the opposite end. By adjustment of the set screws the collimator can be pivoted about the O-ring and brought into alignment with the opposed diamonds and the sample contained therein. In the preferred embodiment there is further provided a set of plugs which are insertable in the bore of the collimator. The plugs have bores of different diameters. By successively inserting plugs of progressively smaller bore diameters and adjusting the alignment of the collimator with each plug, the collimator can be quickly brought into accurate alignment with the diamonds. The collimation system allows alignment of the collimator either before or after the cell has been loaded and pressurized.

  2. Simultaneous reduction of radiation dose and scatter for CBCT by using collimators

    SciTech Connect

    Li, Tianfang; Li, Xiang; Yang, Yong; Zhang, Yongqian; Heron, Dwight E.; Huq, M. Saiful

    2013-12-15

    Purpose: On-board cone-beam CT (CBCT) imaging has been widely available in radiotherapy clinic for target localization. However, the extra radiation dose from CBCT is always a concern for its frequent use. Additionally, the relatively large scatter in CBCT often degrades the image quality. By using collimators, some of the X-rays can be stopped from reaching the patient and the detectors, hence both the scatter and the patient doses are simultaneously reduced. The authors show in this work that the collimated CBCT data can be reconstructed without any noticeable artifacts for certain collimator blocking ratios and blocking patterns, and the focus of this work is to study the relationship between the image quality and these two collimator factors.Methods: A CBCT system with collimators was simulated following the typical geometry used in clinic. Different collimator designs were tested by varying the size and the number of the collimator slits, and at the same time, the ratio of transmitted beams to total beams was varied from 100% to 10%, resulting in hundreds of different simulation scenarios. Lung and pelvis phantoms created from patients CT images were used in the simulations, and an iterative reconstruction algorithm using the compressed sensing technique was adopted. The image quality was examined by root mean square errors (RMSEs) and compared with the conventional CBCT images.Results: The CBCT image quality increases as the amount of beams passing through the collimators increases, and decreases as the size of the collimator slits increases. With ultra-high resolution collimators, the RMSEs were comparable to the conventional CBCT image quality until the beam transmission rate is reduced below 25%.Conclusions: Collimators can reduce the scatters and radiation dose, however, the collimated CBCT image quality is strongly dependent on both the collimator blocking ratio and the blocking pattern. To achieve image quality comparable to the conventional CBCT, the

  3. Fluxes and spectra of quasimonochromatic annihilation photons for studying E1 giant resonances in nuclei

    SciTech Connect

    Dzhilavyan, L. Z.

    2014-12-15

    The fluxes and spectra of quasimonochromatic photons originating from the in-flight annihilation of positrons interacting with electrons of targets are analyzed in the energy region characteristic of the excitation of E1 giant resonances in nuclei. Targets of small thickness and low atomic number are used. The dependences of the spectra on the energy and angle (and their scatter) for positrons incident to the target, on the collimation angle for photons, and on the target thickness are studied.

  4. Near-field collimation of light carrying orbital angular momentum with bull’s-eye-assisted plasmonic coaxial waveguides

    PubMed Central

    Pu, Mingbo; Ma, Xiaoliang; Zhao, Zeyu; Li, Xiong; Wang, Yanqin; Gao, Hui; Hu, Chenggang; Gao, Ping; Wang, Changtao; Luo, Xiangang

    2015-01-01

    The orbital angular momentum (OAM) of light, as an emerging hotspot in optics and photonics, introduces many degrees of freedom for applications ranging from optical communication and quantum processing to micromanipulation. To achieve a high degree of integration, optical circuits for OAM light are essential, which are, however, challenging in the optical regime owing to the lack of well-developed theory. Here we provide a scheme to guide and collimate the OAM beam at the micro- and nano-levels. The coaxial plasmonic slit was exploited as a naturally occurring waveguide for light carrying OAM. Concentric grooves etched on the output surface of the coaxial waveguide were utilized as a plasmonic metasurface to couple the OAM beam to free space with greatly increased beam directivity. Experimental results at λ = 532 nm validated the novel transportation and collimating effect of the OAM beam. Furthermore, dynamic tuning of the topological charges was demonstrated by using a liquid crystal spatial light modulator (SLM). PMID:26159423

  5. Collimated Fast Wind in the Preplanetary Nebula CRL 618

    NASA Astrophysics Data System (ADS)

    Lee, Chin-Fei; Hsu, Ming-Chien; Sahai, Raghvendra

    2009-05-01

    Collimated fast winds (CFWs) have been proposed to operate during the post asymptotic giant branch (post-AGB) evolutionary phase (and even earlier during the late AGB phase), responsible for the shaping of preplanetary nebulae (PPNs) and young planetary nebulae (PNs). This paper is a follow-up to our previous study of CFW models for the well-studied PPN CRL 618. Previously, we compared our CFW models with optical observations of CRL 618 in atomic and ionic lines and found that a CFW with a small opening angle can readily reproduce the highly collimated shape of the northwestern (W1) lobe of CRL 618 and the bowlike structure seen at its tip. In this paper, we compare our CFW models with recent observations of CRL 618 in CO J = 2-1, J = 6-5, and H2 1-0 S(1). In our models, limb-brightened shell structures are seen in CO and H2 at low velocity (LV) arising from the shocked AGB wind in the shell, and can be identified as the LV components in the observations. However, the shell structure in CO J = 2-1 is significantly less extended than that seen in the observations. None of our models can properly reproduce the observed high-velocity (HV) molecular emission near the source along the body of the lobe. In order to reproduce the HV molecular emission in CRL 618, the CFW is required to have a different structure. One possible CFW structure is the cylindrical jet, with the fast wind material confined to a small cross section and collimated to the same direction along the outflow axis.

  6. COLLIMATED FAST WIND IN THE PREPLANETARY NEBULA CRL 618

    SciTech Connect

    Lee, C.-F.; Hsu, M.-C.; Sahai, Raghvendra

    2009-05-10

    Collimated fast winds (CFWs) have been proposed to operate during the post asymptotic giant branch (post-AGB) evolutionary phase (and even earlier during the late AGB phase), responsible for the shaping of preplanetary nebulae (PPNs) and young planetary nebulae (PNs). This paper is a follow-up to our previous study of CFW models for the well-studied PPN CRL 618. Previously, we compared our CFW models with optical observations of CRL 618 in atomic and ionic lines and found that a CFW with a small opening angle can readily reproduce the highly collimated shape of the northwestern (W1) lobe of CRL 618 and the bowlike structure seen at its tip. In this paper, we compare our CFW models with recent observations of CRL 618 in CO J = 2-1, J = 6-5, and H{sub 2} 1-0 S(1). In our models, limb-brightened shell structures are seen in CO and H{sub 2} at low velocity (LV) arising from the shocked AGB wind in the shell, and can be identified as the LV components in the observations. However, the shell structure in CO J = 2-1 is significantly less extended than that seen in the observations. None of our models can properly reproduce the observed high-velocity (HV) molecular emission near the source along the body of the lobe. In order to reproduce the HV molecular emission in CRL 618, the CFW is required to have a different structure. One possible CFW structure is the cylindrical jet, with the fast wind material confined to a small cross section and collimated to the same direction along the outflow axis.

  7. THE FIRST ''WATER FOUNTAIN'' COLLIMATED OUTFLOW IN A PLANETARY NEBULA

    SciTech Connect

    Gómez, José F.; Miranda, Luis F.; Guerrero, Martín A.; Rizzo, J. Ricardo; García-García, Enrique; Green, James A.; Uscanga, Lucero; Ramos-Larios, Gerardo

    2015-02-01

    ''Water fountains'' (WFs) are evolved objects showing high-velocity, collimated jets traced by water maser emission. Most of them are in the post-asymptotic giant branch (post-AGB) and they may represent one of the first manifestations of collimated mass loss in evolved stars. We present water maser, carbon monoxide, and mid-infrared spectroscopic data (obtained with the Australia Telescope Compact Array, Herschel Space Observatory, and the Very Large Telescope, respectively) toward IRAS 15103–5754, a possible planetary nebula (PN) with WF characteristics. Carbon monoxide observations show that IRAS 15103–5754 is an evolved object, while the mid-IR spectrum displays unambiguous [Ne II] emission, indicating that photoionization has started and thus, its nature as a PN is confirmed. Water maser spectra show several components spreading over a large velocity range (≅ 75 km s{sup –1}) and tracing a collimated jet. This indicates that the object is a WF, the first WF known that has already entered the PN phase. However, the spatial and kinematical distribution of the maser emission in this object are significantly different from those in other WFs. Moreover, the velocity distribution of the maser emission shows a ''Hubble-like'' flow (higher velocities at larger distances from the central star), consistent with a short-lived, explosive mass-loss event. This velocity pattern is not seen in other WFs (which are presumably in earlier evolutionary stages). We therefore suggest that we are witnessing a fundamental change of mass-loss processes in WFs, with water masers being pumped by steady jets in post-AGB stars, but tracing explosive/ballistic events as the object enters the PN phase.

  8. Influence of longitudinally tapered collimators on a high brightness electron beam

    NASA Astrophysics Data System (ADS)

    Di Mitri, S.; Froehlich, L.; Karantzoulis, E.

    2012-06-01

    This article presents the design and operation of a longitudinally tapered collimator in a single-pass S-band linac driving a high brightness electron beam. Measurements were done for the transverse emittance growth induced by the collimator wakefield as a function of the lateral displacement of the beam inside the collimator and the energy acceptance provided by an identical collimator installed in a dispersive region. The measurements demonstrate that: (i) the proposed design allows very precise and reproducible motion down to the micron level of the compact, four-hole collimator; (ii) the collimator does not degrade the beam emittance in the presence of standard trajectory control; (iii) the measured kick factor and energy acceptance are in agreement with the theoretical expectations. These measurements were made using 500 pC, 2.4 ps long bunches at the FERMI@Elettra free electron laser facility.

  9. A combined radial collimator and cooled beryllium filter for neutron scattering

    NASA Astrophysics Data System (ADS)

    Groitl, Felix; Rantsiou, Emmanouela; Bartkowiak, Marek; Filges, Uwe; Graf, Dieter; Niedermayer, Christof; Rüegg, Christian; Rønnow, Henrik M.

    2016-05-01

    A flexible, combined, radial collimator and beryllium (Be) filter have been designed and manufactured at the Paul Scherrer Institut (PSI), Switzerland. The Be is integrated in the radial collimator by placing thin Be slices between the collimator lamellas. The filter/collimator is mounted within a vacuum vessel and dry cooled. The flexible design allows for different degrees of collimation and for different Be lengths. Results of measurements carried out at the BOA beamline at PSI are presented. These experiments include rotation scans determining the focal full width half maximum (FWHM), transmission measurements, test of different collimator lamellas and performance tests of the cooling of the filter. This new combined device will be a crucial part of the CAMEA spectrometer at SINQ, PSI.

  10. Plume collimation for laser ablation electrospray ionization mass spectrometry

    DOEpatents

    Vertes, Akos; Stolee, Jessica A.

    2016-06-07

    In various embodiments, a device may generally comprise a capillary having a first end and a second end; a laser to emit energy at a sample in the capillary to ablate the sample and generate an ablation plume in the capillary; an electrospray apparatus to generate an electrospray plume to intercept the ablation plume to produce ions; and a mass spectrometer having an ion transfer inlet to capture the ions. The ablation plume may comprise a collimated ablation plume. The device may comprise a flow cytometer. Methods of making and using the same are also described.

  11. T Pyxidis: The First Cataclysmic Variable with a Collimated Jet

    NASA Technical Reports Server (NTRS)

    Shahbaz, T.; Livio, M.; Southwell, K. A.; Charles, P. A.

    1997-01-01

    We present the first observational evidence for a collimated jet in a cataclysmic variable system; the recurrent nova T Pyxidis. Optical spectra show bipolar components of H(alpha) with velocities approx. 1400 km/s, very similar to those observed in the supersoft X-ray sources and in SS 433. We argue that a key ingredient of the formation of jets in the supersoft X-ray sources and T Pyx (in addition to an accretion disk threaded by a vertical magnetic field), is the presence of nuclear burning on the surface of the white dwarf.

  12. Acceleration and Collimation of Relativistic Magnetohydrodynamic Disk Winds

    NASA Astrophysics Data System (ADS)

    Porth, Oliver; Fendt, Christian

    2010-02-01

    We perform axisymmetric relativistic magnetohydrodynamic simulations to investigate the acceleration and collimation of jets and outflows from disks around compact objects. Newtonian gravity is added to the relativistic treatment in order to establish the physical boundary condition of an underlying accretion disk in centrifugal and pressure equilibrium. The fiducial disk surface (respectively a slow disk wind) is prescribed as boundary condition for the outflow. We apply this technique for the first time in the context of relativistic jets. The strength of this approach is that it allows us to run a parameter study in order to investigate how the accretion disk conditions govern the outflow formation. Substantial effort has been made to implement a current-free, numerical outflow boundary condition in order to avoid artificial collimation present in the standard outflow conditions. Our simulations using the PLUTO code run for 500 inner disk rotations and on a physical grid size of 100 × 200 inner disk radii. The simulations evolve from an initial state in hydrostatic equilibrium and an initially force-free magnetic field configuration. Two options for the initial field geometries are applied—an hourglass-shaped potential magnetic field and a split monopole field. Most of our parameter runs evolve into a steady state solution which can be further analyzed concerning the physical mechanism at work. In general, we obtain collimated beams of mildly relativistic speed with Lorentz factors up to 6 and mass-weighted half-opening angles of 3-7 deg. The split-monopole initial setup usually results in less collimated outflows. The light surface of the outflow magnetosphere tends to align vertically—implying three relativistically distinct regimes in the flow—an inner subrelativistic domain close to the jet axis, a (rather narrow) relativistic jet and a surrounding subrelativistic outflow launched from the outer disk surface—similar to the spine-sheath structure

  13. Design and experimental validation of a compact collimated Knudsen source.

    PubMed

    Wouters, Steinar H W; Ten Haaf, Gijs; Mutsaers, Peter H A; Vredenbregt, Edgar J D

    2016-08-01

    In this paper, the design and performance of a collimated Knudsen source, which has the benefit of a simple design over recirculating sources, is discussed. Measurements of the flux, transverse velocity distribution, and brightness of the resulting rubidium beam at different source temperatures were conducted to evaluate the performance. The scaling of the flux and brightness with the source temperature follows the theoretical predictions. The transverse velocity distribution in the transparent operation regime also agrees with the simulated data. The source was tested up to a temperature of 433 K and was able to produce a flux in excess of 10(13) s(-1). PMID:27587111

  14. Characteristic of laser diode beam propagation through a collimating lens.

    PubMed

    Xu, Qiang; Han, Yiping; Cui, Zhiwei

    2010-01-20

    A mathematical model of a laser diode beam propagating through a collimating lens is presented. Wave propagation beyond the paraxial approximation is studied. The phase delay of the laser diode wave in passing through the lens is analyzed in detail. The propagation optical field after the lens is obtained from the diffraction integral by the stationary phase method. The model is employed to predict the light intensity at various beam cross sections, and the computed intensity distributions are in a good agreement with the corresponding measurements.

  15. Plume collimation for laser ablation electrospray ionization mass spectrometry

    SciTech Connect

    Vertes, Akos; Stolee, Jessica A.

    2014-09-09

    In various embodiments, a device may generally comprise a capillary having a first end and a second end; a laser to emit energy at a sample in the capillary to ablate the sample and generate an ablation plume in the capillary; an electrospray apparatus to generate an electrospray plume to intercept the ablation plume to produce ions; and a mass spectrometer having an ion transfer inlet to capture the ions. The ablation plume may comprise a collimated ablation plume. The device may comprise a flow cytometer. Methods of making and using the same are also described.

  16. ACCELERATION AND COLLIMATION OF RELATIVISTIC MAGNETOHYDRODYNAMIC DISK WINDS

    SciTech Connect

    Porth, Oliver; Fendt, Christian E-mail: fendt@mpia.d

    2010-02-01

    We perform axisymmetric relativistic magnetohydrodynamic simulations to investigate the acceleration and collimation of jets and outflows from disks around compact objects. Newtonian gravity is added to the relativistic treatment in order to establish the physical boundary condition of an underlying accretion disk in centrifugal and pressure equilibrium. The fiducial disk surface (respectively a slow disk wind) is prescribed as boundary condition for the outflow. We apply this technique for the first time in the context of relativistic jets. The strength of this approach is that it allows us to run a parameter study in order to investigate how the accretion disk conditions govern the outflow formation. Substantial effort has been made to implement a current-free, numerical outflow boundary condition in order to avoid artificial collimation present in the standard outflow conditions. Our simulations using the PLUTO code run for 500 inner disk rotations and on a physical grid size of 100 x 200 inner disk radii. The simulations evolve from an initial state in hydrostatic equilibrium and an initially force-free magnetic field configuration. Two options for the initial field geometries are applied-an hourglass-shaped potential magnetic field and a split monopole field. Most of our parameter runs evolve into a steady state solution which can be further analyzed concerning the physical mechanism at work. In general, we obtain collimated beams of mildly relativistic speed with Lorentz factors up to 6 and mass-weighted half-opening angles of 3-7 deg. The split-monopole initial setup usually results in less collimated outflows. The light surface of the outflow magnetosphere tends to align vertically-implying three relativistically distinct regimes in the flow-an inner subrelativistic domain close to the jet axis, a (rather narrow) relativistic jet and a surrounding subrelativistic outflow launched from the outer disk surface-similar to the spine-sheath structure currently

  17. Compact collimated fiber optic array diagnostic for railgun plasma experiments

    SciTech Connect

    Tang, V; Solberg, J; Ferriera, T; Tully, L; Stephan, P

    2008-10-02

    We have developed and tested a compact collimated sixteen channel fiber optic array diagnostic for studying the light emission of railgun armature plasmas with {approx}mm spatial and sub-{micro}s temporal resolution. The design and operational details of the diagnostic are described. Plasma velocities, oscillation, and dimension data from the diagnostic for the Livermore Fixed Hybrid Armature experiment are presented and compared with 1-D simulations. The techniques and principles discussed allow the extension of the diagnostic to other railgun and related dense plasma experiments.

  18. Versatile collimating crystal stage for Bonse-Hart USAXS instrument.

    SciTech Connect

    Ilavsky, J.; Shu, D.; Jemian, P. R.; Long, G. G.

    2007-01-01

    An advanced ultra-small-angle X-ray scattering (USAXS) instrument, using the Bonse-Hart design and installed at APS, is a robust and reliable instrument, providing a scattering vector (q) range of nearly 4 decades (0.00015 to 1 {angstrom}{sup -1}), an intensity dynamic range of up to 9 decades, standard-less absolute intensity calibration, and USAXS imaging capabilities. This type of instrument typically uses channel-cut crystals in both the collimating (before sample) and analyzing (after sample) stages. The optical surfaces of these crystals are finished by etching processes, which leave an orange-peel surface texture, which would compromise the USAXS imaging quality. Therefore optics with highly polished surfaces using separated crystals in both collimating and analyzing stages were developed. A novel design of the optics and mechanical stage uses a fixed gap between the two separated collimating crystals in which a triangular section of the first crystal is removed, allowing for a variable number (1, 2, 4, 6, or 8) of crystal reflections for X-ray energies between 7 and 19 keV. The number of reflections is selected by lateral translation of the collimating crystal pair. Rotational alignment of the second crystal in the pair by an artificial channel-cut crystal mechanism, implemented with a novel high-stiffness weak link actuated by both a picomotor and a piezo-electric transducer, provides the capability to align or adjust an assembly of crystals to achieve the same performance as a single channel-cut crystal with integral weak link. The arrangement of both crystals is held on a removable base that can be remounted with precision within the Si(111) rocking curve on a three-point kinematic mount. Additional tilt adjustments are also provided for initial alignment. This monochromator has proven to be highly robust with respect to motions and vibrations, as well as flexible with respect to selection of number of reflections, and its performance directly resulted

  19. Thallium-201 myocardial tomography with a rotating slant-hole collimator and a large number of projections

    SciTech Connect

    Dale, S.; Bone, D. )

    1990-10-01

    Tomographic imaging of the myocardium was performed using a gamma camera and a 30-degree rotating slant-hole collimator to register 64 projections from a restricted-view angle. Section images were reconstructed with a two-dimensional filtered backprojection technique. Performance in terms of resolution, effects of misalignment, and three-dimensional activity distribution was evaluated in phantom studies. In a limited clinical study, ten consecutive patients were imaged both with single-photon emission computed tomography (SPECT) and the method described. In the new method, the camera was orientated in the 30 degrees left anterior oblique position with 15 degrees cranial tilt. Due to the short distance from the myocardium to the camera, resolution within reconstructed section images was high, the noise level was comparatively low, and the mean activity level in the posterior wall was significantly higher (p less than 0.005) than in SPECT.

  20. Prospects for Integrating a Hollow Electron Lens into the LHC Collimation System

    SciTech Connect

    Smith, Jeffrey Claiborne; Assmann, Ralph; Previtali, Valentina; Shiltsev, Vladimir; Valishev, Alexander; /CERN /Fermilab

    2009-08-03

    It has been proposed to use a hollow electron lens with the LHC beam collimation system [1]. The hollow electron beam would be used as a beam scraper and positioned at a closer sigma than the primary collimators to increase the halo particle diffusion rate striking the primaries. In this paper we use multi-turn beam tracking simulations to analyze the effectiveness of such a lens when integrated into the LHC collimation system.

  1. Dynamic and Stagnating Plasma Flow Leading to Magnetic-Flux-Tube Collimation

    SciTech Connect

    You, S.; Yun, G.S.; Bellan, P.M.

    2005-07-22

    Highly collimated, plasma-filled magnetic-flux tubes are frequently observed on galactic, stellar, and laboratory scales. We propose that a single, universal magnetohydrodynamic pumping process explains why such collimated, plasma-filled magnetic-flux tubes are ubiquitous. Experimental evidence from carefully diagnosed laboratory simulations of astrophysical jets confirms this assertion and is reported here. The magnetohydrodynamic process pumps plasma into a magnetic-flux tube and the stagnation of the resulting flow causes this flux tube to become collimated.

  2. Dynamic and stagnating plasma flow leading to magnetic-flux-tube collimation.

    PubMed

    You, S; Yun, G S; Bellan, P M

    2005-07-22

    Highly collimated, plasma-filled magnetic-flux tubes are frequently observed on galactic, stellar, and laboratory scales. We propose that a single, universal magnetohydrodynamic pumping process explains why such collimated, plasma-filled magnetic-flux tubes are ubiquitous. Experimental evidence from carefully diagnosed laboratory simulations of astrophysical jets confirms this assertion and is reported here. The magnetohydrodynamic process pumps plasma into a magnetic-flux tube and the stagnation of the resulting flow causes this flux tube to become collimated.

  3. A condition on the spatial resolution of IR collimators for testing of thermal imaging systems

    NASA Astrophysics Data System (ADS)

    Chrzanowski, Krzysztof; Lee, Hee Chul; Wrona, Wieslaw

    2000-05-01

    A precise condition on the spatial resolution of the IR collimator for testing thermal imaging systems is presented. The condition can be used even if only the spatial resolution of the IR collimator and that of the system under test, measured using popular definitions, are known. It is shown that when the condition is fulfilled, the thermal image degradation caused by the IR collimator is negligible.

  4. Vesicle Photonics

    SciTech Connect

    Vasdekis, Andreas E.; Scott, E. A.; Roke, Sylvie; Hubbell, J. A.; Psaltis, D.

    2013-04-03

    Thin membranes, under appropriate boundary conditions, can self-assemble into vesicles, nanoscale bubbles that encapsulate and hence protect or transport molecular payloads. In this paper, we review the types and applications of light fields interacting with vesicles. By encapsulating light-emitting molecules (e.g. dyes, fluorescent proteins, or quantum dots), vesicles can act as particles and imaging agents. Vesicle imaging can take place also under second harmonic generation from vesicle membrane, as well as employing mass spectrometry. Light fields can also be employed to transport vesicles using optical tweezers (photon momentum) or directly pertrurbe the stability of vesicles and hence trigger the delivery of the encapsulated payload (photon energy).

  5. Photons Revisited

    NASA Astrophysics Data System (ADS)

    Batic, Matej; Begalli, Marcia; Han, Min Cheol; Hauf, Steffen; Hoff, Gabriela; Kim, Chan Hyeong; Kim, Han Sung; Grazia Pia, Maria; Saracco, Paolo; Weidenspointner, Georg

    2014-06-01

    A systematic review of methods and data for the Monte Carlo simulation of photon interactions is in progress: it concerns a wide set of theoretical modeling approaches and data libraries available for this purpose. Models and data libraries are assessed quantitatively with respect to an extensive collection of experimental measurements documented in the literature to determine their accuracy; this evaluation exploits rigorous statistical analysis methods. The computational performance of the associated modeling algorithms is evaluated as well. An overview of the assessment of photon interaction models and results of the experimental validation are presented.

  6. Green photonics

    NASA Astrophysics Data System (ADS)

    Quan, Frederic

    2012-02-01

    Photonics, the broad merger of electronics with the optical sciences, encompasses such a wide swath of technology that its impact is almost universal in our everyday lives. This is a broad overview of some aspects of the industry and their contribution to the ‘green’ or environmental movement. The rationale for energy conservation is briefly discussed and the impact of photonics on our everyday lives and certain industries is described. Some opinions from industry are presented along with market estimates. References are provided to some of the most recent research in these areas.

  7. Measurement of the effect of collimator generated wakefields on the beams in the SLC

    SciTech Connect

    Bane, K.L.F.; Adolphsen, C.; Decker, F.J.; Emma, P.; Krejcik, P.; Zimmermann, F.

    1995-05-01

    Collimators with adjustable jaws are used in the SLC linac, arcs, and final focus to eliminate the tails of the beams that produce backgrounds in the SLD detectors. However, if the beams are not centered within the jaws of the collimators, transverse wakefields are generated which act to increase the beam emittances. The sensitivity to the beam offset is largest in the vertical collimators in the linac where the small beam sizes (< 100 {mu}m) require that the gap between the jaws be reduced to about a millimeter for effective background reduction. To study the wakefield effect of the collimators, measurements of the induced mean angular kick were made as function of the beam offset in one of the SLC collimators. The authors are interested in both the linear (dipole) behavior near the axis of the collimator and the non-linear behavior near the jaws of the collimator. In this paper they present the results together with a comparison to theoretical predictions. Besides helping thems to quantify the effect of the collimators in the SLC these results are also useful in understanding their effect in future linear colliders, in which collimators will also be important components.

  8. An inversion formula for the exponential Radon transform in spatial domain with variable focal-length fan-beam collimation geometry

    SciTech Connect

    Wen Junhai; Liang Zhengrong

    2006-03-15

    Inverting the exponential Radon transform has a potential use for SPECT (single photon emission computed tomography) imaging in cases where a uniform attenuation can be approximated, such as in brain and abdominal imaging. Tretiak and Metz derived in the frequency domain an explicit inversion formula for the exponential Radon transform in two dimensions for parallel-beam collimator geometry. Progress has been made to extend the inversion formula for fan-beam and varying focal-length fan-beam (VFF) collimator geometries. These previous fan-beam and VFF inversion formulas require a spatially variant filtering operation, which complicates the implementation and imposes a heavy computing burden. In this paper, we present an explicit inversion formula, in which a spatially invariant filter is involved. The formula is derived and implemented in the spatial domain for VFF geometry (where parallel-beam and fan-beam geometries are two special cases). Phantom simulations mimicking SPECT studies demonstrate its accuracy in reconstructing the phantom images and efficiency in computation for the considered collimator geometries.

  9. An inversion formula for the exponential Radon transform in spatial domain with variable focal-length fan-beam collimation geometry

    PubMed Central

    Wen, Junhai; Liang, Zhengrong

    2006-01-01

    Inverting the exponential Radon transform has a potential use for SPECT (single photon emission computed tomography) imaging in cases where a uniform attenuation can be approximated, such as in brain and abdominal imaging. Tretiak and Metz derived in the frequency domain an explicit inversion formula for the exponential Radon transform in two dimensions for parallel-beam collimator geometry. Progress has been made to extend the inversion formula for fan-beam and varying focal-length fan-beam (VFF) collimator geometries. These previous fan-beam and VFF inversion formulas require a spatially-variant filtering operation, which complicates the implementation and imposes a heavy computing burden. In this paper, we present an explicit inversion formula, in which a spatially-invariant filter is involved. The formula is derived and implemented in the spatial domain for VFF geometry (where parallel-beam and fan-beam geometries are two special cases). Phantom simulations mimicking SPECT studies demonstrate its accuracy in reconstructing the phantom images and efficiency in computation for the considered collimator geometries. PMID:16878581

  10. Design and wakefield performance of the new SLC collimators

    SciTech Connect

    Decker, F.J.; Bane, K.; Emma, P.

    1996-08-01

    The very small transverse beam sizes of the flat SLC bunches are 100-170 {mu}m in the horizontal and 30-50 {mu}m in the vertical near the end of the SLAC linac. Unexpectedly large transverse Wakefield kicks were observed from the collimators in this region during 1995. Upon inspection, it was found that the 20 {mu}m gold plating had melted and formed a line of spherules along the beam path. To refurbish the collimators, an improved design was required. The challenging task was to find a surface material with better conductivity than the titanium core to reduce resistive wakefields. The material must also be able to sustain the mechanical stress and heating from beam losses without damage. Vanadium was first chosen for ease of coating, but later TiN was used because it is more chemically inert. Recent beam tests measured expected values for geometric Wakefield kicks, but the resistive wall Wakefield kicks were four times larger than calculated.

  11. X-ray microscopy using collimated and focussed synchrotron radiation

    SciTech Connect

    Jones, K.W.; Kwiatek, W.M.; Gordon, B.M.; Hanson, A.L.; Pounds, J.G.; Rivers, M.L.; Sutton, S.R.; Thompson, A.C.; Underwood, J.H.; Giauque, R.D.

    1987-01-01

    X-ray microscopy is a field that has developed rapidly in recent years. Two different approaches have been used. Zone plates have been employed to produce focused beams with sizes as low as 0.07 ..mu..m for x-ray energies below 1 keV. Images of biological materials and elemental maps for major and minor low Z have been produced using above and below absorption edge differences. At higher energies collimators and focusing mirrors have been used to make small diameter beams for excitation of characteristic K- or L-x rays of all elements in the periodic table. The practicality of a single instrument combining all the features of these two approaches is unclear. The use of high-energy x rays for x-ray microscopy has intrinsic value for characterization of thick samples and determination of trace amounts of most elements. A summary of work done on the X-26 beam line at the National Synchrotron Light Source (NSLS) with collimated and focused x rays with energies above 4 keV is given here. 6 refs., 5 figs., 1 tab.

  12. The Transport of Photons through Light Guides

    NASA Astrophysics Data System (ADS)

    Mammei, Russell

    2000-10-01

    We have designed and optimized the light collection a set of light guides, which couple the scintillator to the photomultiplier tubes (PMTs) of an instrumented collimator. This collimator will be used with the experiments employing linearly-polarized photons in Hall B of Jefferson Lab in the spring of 2001. Using the Monte-Carlo simulation package, Guidem V1.0, we were able to simulate a scintillator-light guide-PMT construction. The scintillators were simulated to emit photons isotropically through out the whole volume of the scintillator. Parameters varied in the simulation were: radius of curvature of the bend, wrapping reflectivities, attenuation length, ripple angle(smoothness of the light guide surface), metal crumple angle(crumple of the aluminum wrapping), and the optical glue interfaces. Also investigated were the indices of refraction for the scintillator-light guide interface as well as the light guide-PMT interface. Given a bent geometry, our studies concluded that a curved elbow shape, 6 cm in length, with a radius of curvature of at least 1.4 cm, will produce a transmission efficiency between 54-55% with reference to the total light created in the scintillator. These light guides should have a ripple angle between 0^circ-3^circ and a PMT index of refraction between 1.3-1.5. The light guide should be wrapped in aluminum and the scintillator should be wrapped a paper with reflectivity around 0.1 to prevent smudging of the plastic.

  13. Photon-photon collisions via relativisitic mirrors

    SciTech Connect

    Koga, James K.

    2012-07-11

    Photon-photon scattering at low energies has been predicted theoretically for many years. However, due to the extremely small cross section there has been no experimental confirmation of this. Due to the rapid increase in laser irradiances and projected peak irradiances in planned facilities regimes could be reached where photon-photon scattering could be experimentally observed. We will first review basic aspects of photon-photon collisions concentrating on the calculation of the photon-photon scattering cross section. Then we will discuss the possibilities for observing these phenomena in ultra-high irradiance laser-plasma interactions involving relativistic mirrors.

  14. Multi-resolution multi-sensitivity design for parallel-hole SPECT collimators.

    PubMed

    Li, Yanzhao; Xiao, Peng; Zhu, Xiaohua; Xie, Qingguo

    2016-07-21

    Multi-resolution multi-sensitivity (MRMS) collimator offering adjustable trade-off between resolution and sensitivity, can make a SPECT system adaptive. We propose in this paper a new idea for MRMS design based on, for the first time, parallel-hole collimators for clinical SPECT. Multiple collimation states with varied resolution/sensitivity trade-offs can be formed by slightly changing the collimator's inner structure. To validate the idea, the GE LEHR collimator is selected as the design prototype and is modeled using a ray-tracing technique. Point images are generated for several states of the design. Results show that the collimation states of the design can obtain similar point response characteristics to parallel-hole collimators, and can be used just like parallel-hole collimators in clinical SPECT imaging. Ray-tracing modeling also shows that the proposed design can offer varied resolution/sensitivity trade-offs: at 100 mm before the collimator, the highest resolution state provides 6.9 mm full width at a half maximum (FWHM) with a nearly minimum sensitivity of about 96.2 cps MBq(-1), while the lowest resolution state obtains 10.6 mm FWHM with the highest sensitivity of about 167.6 cps MBq(-1). Further comparisons of the states on image qualities are conducted through Monte Carlo simulation of a hot-spot phantom which contains five hot spots with varied sizes. Contrast-to-noise ratios (CNR) of the spots are calculated and compared, showing that different spots can prefer different collimation states: the larger spots obtain better CNRs by using the larger sensitivity states, and the smaller spots prefer the higher resolution states. In conclusion, the proposed idea can be an effective approach for MRMS design for parallel-hole SPECT collimators. PMID:27359049

  15. Multi-resolution multi-sensitivity design for parallel-hole SPECT collimators.

    PubMed

    Li, Yanzhao; Xiao, Peng; Zhu, Xiaohua; Xie, Qingguo

    2016-07-21

    Multi-resolution multi-sensitivity (MRMS) collimator offering adjustable trade-off between resolution and sensitivity, can make a SPECT system adaptive. We propose in this paper a new idea for MRMS design based on, for the first time, parallel-hole collimators for clinical SPECT. Multiple collimation states with varied resolution/sensitivity trade-offs can be formed by slightly changing the collimator's inner structure. To validate the idea, the GE LEHR collimator is selected as the design prototype and is modeled using a ray-tracing technique. Point images are generated for several states of the design. Results show that the collimation states of the design can obtain similar point response characteristics to parallel-hole collimators, and can be used just like parallel-hole collimators in clinical SPECT imaging. Ray-tracing modeling also shows that the proposed design can offer varied resolution/sensitivity trade-offs: at 100 mm before the collimator, the highest resolution state provides 6.9 mm full width at a half maximum (FWHM) with a nearly minimum sensitivity of about 96.2 cps MBq(-1), while the lowest resolution state obtains 10.6 mm FWHM with the highest sensitivity of about 167.6 cps MBq(-1). Further comparisons of the states on image qualities are conducted through Monte Carlo simulation of a hot-spot phantom which contains five hot spots with varied sizes. Contrast-to-noise ratios (CNR) of the spots are calculated and compared, showing that different spots can prefer different collimation states: the larger spots obtain better CNRs by using the larger sensitivity states, and the smaller spots prefer the higher resolution states. In conclusion, the proposed idea can be an effective approach for MRMS design for parallel-hole SPECT collimators.

  16. Photon Collider Physics with Real Photon Beams

    SciTech Connect

    Gronberg, J; Asztalos, S

    2005-11-03

    Photon-photon interactions have been an important probe into fundamental particle physics. Until recently, the only way to produce photon-photon collisions was parasitically in the collision of charged particles. Recent advances in short-pulse laser technology have made it possible to consider producing high intensity, tightly focused beams of real photons through Compton scattering. A linear e{sup +}e{sup -} collider could thus be transformed into a photon-photon collider with the addition of high power lasers. In this paper they show that it is possible to make a competitive photon-photon collider experiment using the currently mothballed Stanford Linear Collider. This would produce photon-photon collisions in the GeV energy range which would allow the discovery and study of exotic heavy mesons with spin states of zero and two.

  17. Using Rose’s metal alloy as a pinhole collimator material in preclinical small-animal imaging: A Monte Carlo evaluation

    SciTech Connect

    Peterson, Mikael Strand, Sven-Erik; Ljungberg, Michael

    2015-04-15

    Purpose: Pinhole collimation is the most common method of high-resolution preclinical single photon emission computed tomography imaging. The collimators are usually constructed from dense materials with high atomic numbers, such as gold and platinum, which are expensive and not always flexible in the fabrication step. In this work, the authors have investigated the properties of a fusible alloy called Rose’s metal and its potential in pinhole preclinical imaging. When compared to current standard pinhole materials such as gold and platinum, Rose’s metal has a lower density and a relatively low effective atomic number. However, it is inexpensive, has a low melting point, and does not contract when solidifying. Once cast, the piece can be machined with high precision. The aim of this study was to evaluate the imaging properties for Rose’s metal and compare them with those of standard materials. Methods: After validating their Monte Carlo code by comparing its results with published data and the results from analytical calculations, they investigated different pinhole geometries by varying the collimator material, acceptance angle, aperture diameter, and photon incident angle. The penetration-to-scatter and penetration-to-total component ratios, sensitivity, and the spatial resolution were determined for gold, tungsten, and Rose’s metal for two radionuclides, {sup 99}Tc{sup m} and {sup 125}I. Results: The Rose’s metal pinhole-imaging simulations show higher penetration/total and scatter/total ratios. For example, the penetration/total is 50% for gold and 75% for Rose’s metal when simulating {sup 99}Tc{sup m} with a 0.3 mm aperture diameter and a 60° acceptance angle. However, the degradation in spatial resolution remained below 10% relative to the spatial resolution for gold for acceptance angles below 40° and aperture diameters larger than 0.5 mm. Conclusions: Extra penetration and scatter associated with Rose’s metal contribute to degradation in the

  18. Microalgae photonics

    NASA Astrophysics Data System (ADS)

    Floume, Timmy; Coquil, Thomas; Sylvestre, Julien

    2011-05-01

    Due to their metabolic flexibility and fast growth rate, microscopic aquatic phototrophs like algae have a potential to become industrial photochemical converters. Algae photosynthesis could enable the large scale production of clean and renewable liquid fuels and chemicals with major environmental, economic and societal benefits. Capital and operational costs are the main issues to address through optical, process and biochemical engineering improvements. In this perspective, a variety of photonic approaches have been proposed - we introduce them here and describe their potential, limitations and compatibility with separate biotechnology and engineering progresses. We show that only sunlight-based approaches are economically realistic. One of photonics' main goals in the algae field is to dilute light to overcome photosaturation effects that impact upon cultures exposed to full sunlight. Among other approaches, we introduce a widely-compatible broadband spectral adaptation technique called AlgoSun® that uses luminescence to optimize sunlight spectrum in view of the bioconverter's requirements.

  19. Photon detectors

    SciTech Connect

    Va`vra, J.

    1995-10-01

    J. Seguinot and T. Ypsilantis have recently described the theory and history of Ring Imaging Cherenkov (RICH) detectors. In this paper, I will expand on these excellent review papers, by covering the various photon detector designs in greater detail, and by including discussion of mistakes made, and detector problems encountered, along the way. Photon detectors are among the most difficult devices used in physics experiments, because they must achieve high efficiency for photon transport and for the detection of single photo-electrons. For gaseous devices, this requires the correct choice of gas gain in order to prevent breakdown and wire aging, together with the use of low noise electronics having the maximum possible amplification. In addition, the detector must be constructed of materials which resist corrosion due to photosensitive materials such as, the detector enclosure must be tightly sealed in order to prevent oxygen leaks, etc. The most critical step is the selection of the photocathode material. Typically, a choice must be made between a solid (CsI) or gaseous photocathode (TMAE, TEA). A conservative approach favors a gaseous photocathode, since it is continuously being replaced by flushing, and permits the photon detectors to be easily serviced (the air sensitive photocathode can be removed at any time). In addition, it can be argued that we now know how to handle TMAE, which, as is generally accepted, is the best photocathode material available as far as quantum efficiency is concerned. However, it is a very fragile molecule, and therefore its use may result in relatively fast wire aging. A possible alternative is TEA, which, in the early days, was rejected because it requires expensive CaF{sub 2} windows, which could be contaminated easily in the region of 8.3 eV and thus lose their UV transmission.

  20. Adaptive fiber optics collimator based on flexible hinges.

    PubMed

    Zhi, Dong; Ma, Yanxing; Ma, Pengfei; Si, Lei; Wang, Xiaolin; Zhou, Pu

    2014-08-20

    In this manuscript, we present a new design for an adaptive fiber optics collimator (AFOC) based on flexible hinges by using piezoelectric stacks actuators for X-Y displacement. Different from traditional AFOC, the new structure is based on flexible hinges to drive the fiber end cap instead of naked fiber. We fabricated a real AFOC based on flexible hinges, and the end cap's deviation and resonance frequency of the device were measured. Experimental results show that this new AFOC can provide fast control of tip-tilt deviation of the laser beam emitting from the end cap. As a result, the fiber end cap can support much higher power than naked fiber, which makes the new structure ideal for tip-tilt controlling in a high-power fiber laser system.

  1. Performance evaluation of advanced industrial SPECT system with diverging collimator.

    PubMed

    Park, Jang Guen; Jung, Sung-Hee; Kim, Jong Bum; Moon, Jinho; Yeom, Yeon Soo; Kim, Chan Hyeong

    2014-12-01

    An advanced industrial SPECT system with 12-fold-array diverging collimator was developed for flow visualization in industrial reactors and was discussed in the previous study. The present paper describes performance evaluation of the SPECT system under both static- and dynamic- flow conditions. Under static conditions, the movement of radiotracer inside the test reactor was compared with that of color tracer (blue ink) captured with a high-speed camera. The comparison of the reconstructed images obtained with the radiotracer and the SPECT system showed fairly good agreement with video-frames of the color tracer obtained with the camera. Based on the results of the performance evaluation, it is concluded that the SPECT system is suitable for investigation and visualization of flows in industrial flow reactors.

  2. Simulations of the Fermilab Recycler for Losses and Collimation

    SciTech Connect

    Stern, Eric; Ainsworth, Robert; Amundson, James; Brown, Bruce

    2015-06-01

    Fermilab has recently completed an upgrade to the com- plex with the goal of delivering 700 kW of beam power as 120 GeV protons to the NuMI target. A major part of boost- ing beam power is to shorten the beam cycle by accumulating up to 12 bunches of 0.5 × 10 11 protons in the Recycler ring through slip-stacking during the Main Injector ramp. This introduces much higher intensities into the Recycler than it has had before. Meeting radiation safety requirements with high intensity operations requires understanding the ef- fects of space charge induced tune spreads and resulting halo formation, and aperture restrictions in the real machine to de- velop a collimation strategy. We report on initial simulations of slip-stacking in the Recycler performed with Synergia.

  3. Approaching Collimation with a Graphene-based Quantum Point Contact

    NASA Astrophysics Data System (ADS)

    Pan, Grace; Lee, Menyoung; Watanabe, Kenji; Taniguchi, Takashi; Goldhaber-Gordon, David

    Quantum point contacts (QPCs) are narrow constrictions on the order of the Fermi wavelength that bridge together two electrically conducting regions. QPCs display sensitive conductance quantization and are a classic playing field to illustrate clean, ballistic transport in low-dimensional materials. However, graphene-based QPCs are challenging to fabricate, in part due to two reasons: edge disorder that suppresses conductance quantization and imperfect gate depletion leading to charge puddles. Using graphene-boron nitride heterostructures, we demonstrate improvements over a simple etch and Au-gating method by introducing a protective alumina dielectric layer. We use this method to create two QPCs in series and explore potential electron-beam collimation at low magnetic field, in the spirit of Molenkamp (1990).

  4. Solar wind collimation of the Jupiter high velocity dust streams

    NASA Astrophysics Data System (ADS)

    Flandes, A.; Krueger, H.

    2006-12-01

    The dust bursts discovered by the Ulysses dust sensor when approaching Jupiter in 1992 were later confirmed as collimated streams of high velocity (~200 km/s) charged (~5V) dust grains escaping from Jupiter and dominated by the interplanetary Magnetic field (IMF). With Cassini, a similar phenomenon was observed in Saturn. It was demonstrated that the Jovian dust streams are closely related to the solar wind compressed regions, either Corotating interaction regions (CIRs) or Coronal mass ejections (CMEs) ¨Cto a minor extent-. Actually the dust streams seem ultimately to be generated by such events. This can be explained considering that dust grains are accelerated as they gain substantial energy while compressed at the forward and reverse shocks that bound or precede these solar wind regions.

  5. Performance evaluation of advanced industrial SPECT system with diverging collimator.

    PubMed

    Park, Jang Guen; Jung, Sung-Hee; Kim, Jong Bum; Moon, Jinho; Yeom, Yeon Soo; Kim, Chan Hyeong

    2014-12-01

    An advanced industrial SPECT system with 12-fold-array diverging collimator was developed for flow visualization in industrial reactors and was discussed in the previous study. The present paper describes performance evaluation of the SPECT system under both static- and dynamic- flow conditions. Under static conditions, the movement of radiotracer inside the test reactor was compared with that of color tracer (blue ink) captured with a high-speed camera. The comparison of the reconstructed images obtained with the radiotracer and the SPECT system showed fairly good agreement with video-frames of the color tracer obtained with the camera. Based on the results of the performance evaluation, it is concluded that the SPECT system is suitable for investigation and visualization of flows in industrial flow reactors. PMID:25169132

  6. Note: Detector collimators for the nanoscale ordered materials diffractometer instrument at the Spallation Neutron Source.

    PubMed

    Tamalonis, A; Weber, J K R; Neuefeind, J C; Carruth, J; Skinner, L B; Alderman, O L G; Benmore, C J

    2015-09-01

    Five neutron collimator designs were constructed and tested at the nanoscale ordered materials diffractometer (NOMAD) instrument. Collimators were made from High Density PolyEthylene (HDPE) or 5% borated HDPE. In all cases, collimators improved the signal to background ratio and reduced detection of secondary scattering. In the Q-range 10-20 Å(-1), signal to background ratio improved by factors of approximately 1.6 and 2.0 for 50 and 100 mm deep collimators, respectively. In the Q-range 40-50 Å(-1), the improvement factors were 1.8 and 2.7. Secondary scattering as measured at Q ∼ 9.5 Å(-1) was significantly decreased when the collimators were installed. PMID:26429492

  7. A new CT collimator for producing two simultaneous overlapping slices from one scan. [for biomedical applications

    NASA Technical Reports Server (NTRS)

    Kwoh, Y. S.; Glenn, W. V., Jr.; Reed, I. S.; Truong, T. K.

    1981-01-01

    A new CT collimator is developed which is capable of producing two simultaneous successive overlapping images from a single scan. The collimator represents a modification of the standard EMI 5005 collimator achieved by alternately masking one end or portions of both ends of the X-ray detectors at a 13-mm beamwidth so that a set of 540 filtered projections is obtained for each scan which can be separated into two sets of interleaved projections corresponding to views 3 mm apart. Tests have demonstrated that the quality of the images produced from these two projections almost equals the quality of those produced by the standard collimator from two separate scans. The new collimator may thus be used to achieve a speed improvement in the generation of overlapping sections as well as a reduction in X-ray dosage.

  8. Note: Detector collimators for the nanoscale ordered materials diffractometer instrument at the Spallation Neutron Source

    SciTech Connect

    Tamalonis, A.; Weber, J. K. R.; Neuefeind, J. C.; Carruth, J.; Skinner, L. B.; Alderman, O. L. G.; Benmore, C. J.

    2015-09-01

    Five neutron collimator designs were constructed and tested at the nanoscale ordered materials diffractometer (NOMAD) instrument. Collimators were made from High Density PolyEthylene (HDPE) or 5% borated HDPE. In all cases, collimators improved the signal to background ratio and reduced detection of secondary scattering. In the Q-range 10-20 (angstrom)-1, signal to background ratio improved by factors of approximately 1.6 and 2.0 for 50 and 100 mm deep collimators, respectively. In the Q-range 40-50 angstrom-1, the improvement factors were 1.8 and 2.7. Secondary scattering as measured at Q similar to 9.5 angstrom-1 was significantly decreased when the collimators were installed.

  9. Note: Detector collimators for the nanoscale ordered materials diffractometer instrument at the Spallation Neutron Source

    SciTech Connect

    Tamalonis, A.; Weber, J. K. R. Alderman, O. L. G.; Neuefeind, J. C.; Carruth, J.; Skinner, L. B.; Benmore, C. J.

    2015-09-15

    Five neutron collimator designs were constructed and tested at the nanoscale ordered materials diffractometer (NOMAD) instrument. Collimators were made from High Density PolyEthylene (HDPE) or 5% borated HDPE. In all cases, collimators improved the signal to background ratio and reduced detection of secondary scattering. In the Q-range 10-20 Å{sup −1}, signal to background ratio improved by factors of approximately 1.6 and 2.0 for 50 and 100 mm deep collimators, respectively. In the Q-range 40-50 Å{sup −1}, the improvement factors were 1.8 and 2.7. Secondary scattering as measured at Q ∼ 9.5 Å{sup −1} was significantly decreased when the collimators were installed.

  10. Note: Detector collimators for the nanoscale ordered materials diffractometer instrument at the Spallation Neutron Source

    SciTech Connect

    Tamalonis, A.; Weber, J. K. R.; Neuefeind, J. C.; Carruth, J.; Skinner, L. B.; Alderman, O. L. G.; Benmore, C. J.

    2015-09-09

    We constructed and tested five neutron collimator designs using the nanoscale ordered materials diffractometer (NOMAD) instrument. Collimators were made from High Density PolyEthylene (HDPE) or 5% borated HDPE. In all cases, collimators improved the signal to background ratio and reduced detection of secondary scattering. Moreover, in the Q-range 10-20 Å-1, signal to background ratio improved by factors of approximately 1.6 and 2.0 for 50 and 100 mm deep collimators, respectively. In the Q-range 40-50 Å-1, the improvement factors were 1.8 and 2.7. Secondary scattering as measured at Q similar to 9.5 Å-1 was significantly decreased when the collimators were installed.

  11. Note: Detector collimators for the nanoscale ordered materials diffractometer instrument at the Spallation Neutron Source

    DOE PAGES

    Tamalonis, A.; Weber, J. K. R.; Neuefeind, J. C.; Carruth, J.; Skinner, L. B.; Alderman, O. L. G.; Benmore, C. J.

    2015-09-09

    We constructed and tested five neutron collimator designs using the nanoscale ordered materials diffractometer (NOMAD) instrument. Collimators were made from High Density PolyEthylene (HDPE) or 5% borated HDPE. In all cases, collimators improved the signal to background ratio and reduced detection of secondary scattering. Moreover, in the Q-range 10-20 Å-1, signal to background ratio improved by factors of approximately 1.6 and 2.0 for 50 and 100 mm deep collimators, respectively. In the Q-range 40-50 Å-1, the improvement factors were 1.8 and 2.7. Secondary scattering as measured at Q similar to 9.5 Å-1 was significantly decreased when the collimators were installed.

  12. Note: Detector collimators for the nanoscale ordered materials diffractometer instrument at the Spallation Neutron Source

    NASA Astrophysics Data System (ADS)

    Tamalonis, A.; Weber, J. K. R.; Neuefeind, J. C.; Carruth, J.; Skinner, L. B.; Alderman, O. L. G.; Benmore, C. J.

    2015-09-01

    Five neutron collimator designs were constructed and tested at the nanoscale ordered materials diffractometer (NOMAD) instrument. Collimators were made from High Density PolyEthylene (HDPE) or 5% borated HDPE. In all cases, collimators improved the signal to background ratio and reduced detection of secondary scattering. In the Q-range 10-20 Å-1, signal to background ratio improved by factors of approximately 1.6 and 2.0 for 50 and 100 mm deep collimators, respectively. In the Q-range 40-50 Å-1, the improvement factors were 1.8 and 2.7. Secondary scattering as measured at Q ˜ 9.5 Å-1 was significantly decreased when the collimators were installed.

  13. Polarization sensitive beam bending using a spatially variant photonic crystal

    NASA Astrophysics Data System (ADS)

    Digaum, Jennefir L.; Pazos, Javier; Rumpf, Raymond; Chiles, Jeff; Fathpour, Sasan; Thomas, Jeremy N.; Kuebler, Stephen M.

    2015-02-01

    A spatially-variant photonic crystal (SVPC) that can control the spatial propagation of electromagnetic waves in three dimensions with high polarization sensitivity was fabricated and characterized. The geometric attributes of the SVPC lattice were spatially varied to make use of the directional phenomena of self-collimation to tightly bend an unguided beam coherently through a 90 degree angle. Both the lattice spacing and the fill factor of the SVPC were maintained to be nearly constant throughout the structure. A finite-difference frequency-domain computational method confirms that the SVPC can self-collimate and bend light without significant diffuse scatter caused by the bend. The SVPC was fabricated using multi-photon direct laser writing in the photo-polymer SU-8. Mid-infrared light having a vacuum wavelength of λ0 = 2.94 μm was used to experimentally characterize the SVPCs by scanning the sides of the structure with optical fibers and measuring the intensity of light emanating from each face. Results show that the SVPC is capable of directing power flow of one polarization through a 90-degree turn, confirming the self-collimating and polarization selective light-guiding properties of the structures.

  14. High-energy photon radiography system using laser-Compton scattering for inspection of bulk materials

    NASA Astrophysics Data System (ADS)

    Toyokawa, Hiroyuki; Ohgaki, Hideaki; Mikado, Tomohisa; Yamada, Kawakatsu

    2002-09-01

    Transmission radiography of bulk materials using continuously energy-tunable quasimonochromatic photon beams has been performed with the photons of a few tens MeV. The photon beam is generated with the laser-Compton scattering of the 300-800 MeV electrons in the electron storage ring "TERAS" with laser lights. The present radiography system is applicable to an inspection of bulk materials and to nondestructive testing of large industrial products. The effectiveness of the method has been examined using metals, ceramics, and concrete blocks. The spatial resolution of the radiography system was measured to be 650 mum using a 1 mm collimator system.

  15. A review on photoneutrons characteristics in radiation therapy with high-energy photon beams

    PubMed Central

    Naseri, Alireza; Mesbahi, Asghar

    2010-01-01

    In radiation therapy with high-energy photon beams (E > 10 MeV) neutrons are generated mainly in linacs head thorough (γ,n) interactions of photons with nuclei of high atomic number materials that constitute the linac head and the beam collimation system. These neutrons affect the shielding requirements in radiation therapy rooms and also increase the out-of-field radiation dose of patients undergoing radiation therapy with high-energy photon beams. In the current review, the authors describe the factors influencing the neutron production for different medical linacs based on the performed measurements and Monte Carlo studies in the literature. PMID:24376940

  16. Parametrization and application of scatter kernels for modelling scanned proton beam collimator scatter dose

    NASA Astrophysics Data System (ADS)

    Kimstrand, Peter; Traneus, Erik; Ahnesjö, Anders; Tilly, Nina

    2008-07-01

    Collimators are routinely used in proton radiotherapy to laterally confine the field and improve the penumbra. Collimator scatter contributes up to 15% of the local dose and is therefore important to include in treatment planning dose calculation. We present a method for reconstruction of the collimator scatter phase space based on the parametrization of pre-calculated scatter kernels. Collimator scatter distributions, generated by the Monte Carlo (MC) package GEANT4.8.2, were scored differential in direction and energy. The distributions were then parametrized so as to enable a fast reconstruction by sampling. MC calculated dose distributions in water based on the parametrized phase space were compared to full MC simulations that included the collimator in the simulation geometry, as well as to experimental data. The experiments were performed at the scanned proton beam line at the The Svedberg Laboratory (TSL) in Uppsala, Sweden. Dose calculations using the parametrization of this work and the full MC for isolated typical cases of collimator scatter were compared by means of the gamma index. The result showed that in total 96.7% (99.3%) of the voxels fulfilled the gamma 2.0%/2.0 mm (3.0%/3.0 mm) criterion. The dose distribution for a collimated field was calculated based on the phase space created by the collimator scatter model incorporated into the generation of the phase space of a scanned proton beam. Comparing these dose distributions to full MC simulations, including particle transport in the MLC, yielded that in total for 18 different collimated fields, 99.1% of the voxels satisfied the gamma 1.0%/1.0 mm criterion and no voxel exceeded the gamma 2.6%/2.6 mm criterion. The dose contribution of collimator scatter along the central axis as predicted by the model showed good agreement with experimental data.

  17. Multi-resolution multi-sensitivity design for parallel-hole SPECT collimators

    NASA Astrophysics Data System (ADS)

    Li, Yanzhao; Xiao, Peng; Zhu, Xiaohua; Xie, Qingguo

    2016-07-01

    Multi-resolution multi-sensitivity (MRMS) collimator offering adjustable trade-off between resolution and sensitivity, can make a SPECT system adaptive. We propose in this paper a new idea for MRMS design based on, for the first time, parallel-hole collimators for clinical SPECT. Multiple collimation states with varied resolution/sensitivity trade-offs can be formed by slightly changing the collimator’s inner structure. To validate the idea, the GE LEHR collimator is selected as the design prototype and is modeled using a ray-tracing technique. Point images are generated for several states of the design. Results show that the collimation states of the design can obtain similar point response characteristics to parallel-hole collimators, and can be used just like parallel-hole collimators in clinical SPECT imaging. Ray-tracing modeling also shows that the proposed design can offer varied resolution/sensitivity trade-offs: at 100 mm before the collimator, the highest resolution state provides 6.9 mm full width at a half maximum (FWHM) with a nearly minimum sensitivity of about 96.2 cps MBq‑1, while the lowest resolution state obtains 10.6 mm FWHM with the highest sensitivity of about 167.6 cps MBq‑1. Further comparisons of the states on image qualities are conducted through Monte Carlo simulation of a hot-spot phantom which contains five hot spots with varied sizes. Contrast-to-noise ratios (CNR) of the spots are calculated and compared, showing that different spots can prefer different collimation states: the larger spots obtain better CNRs by using the larger sensitivity states, and the smaller spots prefer the higher resolution states. In conclusion, the proposed idea can be an effective approach for MRMS design for parallel-hole SPECT collimators.

  18. An MLC-based version for the ecliptic method for the determination of backscatter into the beam monitor chambers in photon beams of medical accelerators.

    PubMed

    Nelli, Flavio Enrico

    2016-03-01

    A very simple method to measure the effect of the backscatter from secondary collimators into the beam monitor chambers in linear accelerators equipped with multi-leaf collimators (MLC) is presented here. The backscatter to the monitor chambers from the upper jaws of the secondary collimator was measured on three beam-matched linacs by means of three methods: this new methodology, the ecliptic method, and assessing the variation of the beam-on time per monitor unit with dose rate feedback disabled. This new methodology was used to assess the backscatter characteristics of asymmetric over-traveling jaws. Excellent agreement between the backscatter values measured using the new methodology introduced here and the ones obtained using the other two methods was established. The experimental values reported here differ by less than 1% from published data. The sensitivity of this novel technique allowed differences in backscatter due to the same opening of the jaws, when placed at different positions on the beam path, to be resolved. The introduction of the ecliptic method has made the determination of the backscatter to the monitor chambers an easy procedure. The method presented here for machines equipped with MLCs makes the determination of backscatter to the beam monitor chambers even easier, and suitable to characterize linacs equipped with over-traveling asymmetric secondary collimators. This experimental procedure could be simply implemented to fully characterize the backscatter output factor constituent when detailed dosimetric modeling of the machine's head is required. The methodology proved to be uncomplicated, accurate and suitable for clinical or experimental environments. PMID:26671445

  19. Collimator optimization and collimator-detector response compensation in myocardial perfusion SPECT using the ideal observer with and without model mismatch and an anthropomorphic model observer

    NASA Astrophysics Data System (ADS)

    Ghaly, Michael; Links, Jonathan M.; Frey, Eric C.

    2016-03-01

    The collimator is the primary factor that determines the spatial resolution and noise tradeoff in myocardial perfusion SPECT images. In this paper, the goal was to find the collimator that optimizes the image quality in terms of a perfusion defect detection task. Since the optimal collimator could depend on the level of approximation of the collimator-detector response (CDR) compensation modeled in reconstruction, we performed this optimization for the cases of modeling the full CDR (including geometric, septal penetration and septal scatter responses), the geometric CDR, or no model of the CDR. We evaluated the performance on the detection task using three model observers. Two observers operated on data in the projection domain: the Ideal Observer (IO) and IO with Model-Mismatch (IO-MM). The third observer was an anthropomorphic Channelized Hotelling Observer (CHO), which operated on reconstructed images. The projection-domain observers have the advantage that they are computationally less intensive. The IO has perfect knowledge of the image formation process, i.e. it has a perfect model of the CDR. The IO-MM takes into account the mismatch between the true (complete and accurate) model and an approximate model, e.g. one that might be used in reconstruction. We evaluated the utility of these projection domain observers in optimizing instrumentation parameters. We investigated a family of 8 parallel-hole collimators, spanning a wide range of resolution and sensitivity tradeoffs, using a population of simulated projection (for the IO and IO-MM) and reconstructed (for the CHO) images that included background variability. We simulated anterolateral and inferior perfusion defects with variable extents and severities. The area under the ROC curve was estimated from the IO, IO-MM, and CHO test statistics and served as the figure-of-merit. The optimal collimator for the IO had a resolution of 9-11 mm FWHM at 10 cm, which is poorer resolution than typical collimators

  20. Development and calibration of fine collimators for the ASTRO-H Soft Gamma-ray Detector

    NASA Astrophysics Data System (ADS)

    Mizuno, T.; Kimura, D.; Fukazawa, Y.; Furui, S.; Goto, K.; Hayashi, T.; Kawabata, K. S.; Kawano, T.; Kitamura, Y.; Shirakawa, H.; Tanabe, T.; Makishima, K.; Nakajima, K.; Nakazawa, K.; Fukuyama, T.; Ichinohe, Y.; Ishimura, K.; Ohta, M.; Sato, T.; Takahashi, T.; Uchida, Y.; Watanabe, S.; Ishibashi, K.; Sakanobe, K.; Matsumoto, H.; Miyazawa, T.; Mori, H.; Sakai, M.; Tajima, H.

    2014-07-01

    The Soft Gamma-ray Detector (SGD) is a Si/CdTe Compton telescope surrounded by a thick BGO active shield and is scheduled to be onboard the ASTRO-H satellite when it is launched in 2015. The SGD covers the energy range from 40 to 600 keV with high sensitivity, which allows us to study nonthermal phenomena in the universe. The SGD uses a Compton camera with the narrow field-of-view (FOV) concept to reduce the non-Xray background (NXB) and improve the sensitivity. Since the SGD is essentially a nonimaging instrument, it also has to cope with the cosmic X-ray background (CXB) within the FOV. The SGD adopts passive shields called "fine collimators" (FCs) to restrict the FOV to <= 0.6° for low-energy photons (<= 100 keV), which reduces contamination from CXB to less than what is expected due to NXB. Although the FC concept was already adopted by the Hard X-ray Detector onboard Suzaku, FCs for the SGD are about four times larger in size and are technically more difficult to operate. We developed FCs for the SGD and confirmed that the prototypes function as required by subjecting them to an X-ray test and environmental tests, such as vibration tests. We also developed an autocollimator system, which uses visible light to determine the transmittance and the optical axis, and calibrated it against data from the X-ray test. The acceptance tests of flight models started in December 2013: five out of six FCs were deemed acceptable, and one more unit is currently being produced. The activation properties were studied based on a proton-beam test and the results were used to estimate the in-orbit NXB.

  1. Photon Calorimeter

    DOEpatents

    Chow, Tze-Show

    1989-01-01

    A photon calorimeter (20, 40) is provided that comprises a laminar substrate (10, 22, 42) that is uniform in density and homogeneous in atomic composition. A plasma-sprayed coating (28, 48, 52), that is generally uniform in density and homogeneous in atomic composition within the proximity of planes that are parallel to the surfaces of the substrate, is applied to either one or both sides of the laminar substrate. The plasma-sprayed coatings may be very efficiently spectrally tailored in atomic number. Thermocouple measuring junctions (30, 50, 54) are positioned within the plasma-sprayed coatings. The calorimeter is rugged, inexpensive, and equilibrates in temperature very rapidly.

  2. Photon calorimeter

    DOEpatents

    Chow, Tze-Show

    1988-04-22

    A photon calorimeter is provided that comprises a laminar substrate that is uniform in density and homogeneous in atomic composition. A plasma-sprayed coating, that is generally uniform in density and homogeneous in atomic composition within the proximity of planes that are parallel to the surfaces of the substrate, is applied to either one or both sides of the laminar substrate. The plasma-sprayed coatings may be very efficiently spectrally tailored in atomic number. Thermocouple measuring junctions, are positioned within the plasma-sprayed coatings. The calorimeter is rugged, inexpensive, and equilibrates in temperature very rapidly. 4 figs.

  3. A method for characterizing photon radiation fields

    SciTech Connect

    Whicker, J.J.; Hsu, H.H.; Hsieh, F.H.; Borak, T.B.

    1999-04-01

    Uncertainty in dosimetric and exposure rate measurements can increase in areas where multi-directional and low-energy photons (< 100 keV) exist because of variations in energy and angular measurement response. Also, accurate measurement of external exposures in spatially non-uniform fields may require multiple dosimetry. Therefore, knowledge of the photon fields in the workplace is required for full understanding of the accuracy of dosimeters and instruments, and for determining the need for multiple dosimeters. This project was designed to develop methods to characterize photon radiation fields in the workplace, and to test the methods in a plutonium facility. The photon field at selected work locations was characterized using TLDs and a collimated NaI(Tl) detector from which spatial variations in photon energy distributions were calculated from measured spectra. Laboratory results showed the accuracy and utility of the method. Field measurement results combined with observed work patterns suggested the following: (1) workers are exposed from all directions, but not isotropically, (2) photon energy distributions were directionally dependent, (3) stuffing nearby gloves into the glovebox reduced exposure rates significantly, (4) dosimeter placement on the front of the chest provided for a reasonable estimate of the average dose equivalent to workers` torsos, (5) justifiable conclusions regarding the need for multiple dosimetry can be made using this quantitative method, and (6) measurements of the exposure rates with ionization chambers pointed with open beta windows toward the glovebox provided the highest measured rates, although absolute accuracy of the field measurements still needs to be assessed.

  4. An investigation of the dose distribution effect related with collimator angle for VMAT method

    NASA Astrophysics Data System (ADS)

    Tas, B.; Bilge, H.; Ozturk, S. Tokdemir

    2016-03-01

    Aim of this study is to investigate the efficacy of dose distribution in eleven prostate cancer patients with single VMAT and double VMAT when varying collimator angle. We generated optimum single and double VMAT treatment plans when collimator angle was 0°. We recalculated single VMAT plans at different collimator angles(0°,15°,30°,45°,60°,75°,90°) for double VMAT plans(0°-0°,15°-345°,30°-330°,45°-315°,60°-300°,75°-285°,90°-270°) without changing any optimization parameters. HI, DVH and %95 dose coverage of PTV calculated and analyzed. We determined better dose distribution with some collimator angles. Plans were verified using the 2 dimensional ion chamber array Matrixx® and 3 dimensional Compass® software program. A higher %95 dose coverage of PTV was found for single VMAT in the 15° collimator angle, for double VMAT in the 60°-300° and 75°-285° collimator angles. Because of lower rectum doses, we suggested 75°-285°. When we compared single and double VMAT's dose distribution, we had better % 95 dose coverage of PTV and lower HI with double VMAT. Our result was significant statistically. These finds are informative for choosing 75°-285° collimator angles in double VMAT plans for prostate cancer.

  5. Beam feasibility study of a collimator with in-jaw beam position monitors

    NASA Astrophysics Data System (ADS)

    Wollmann, Daniel; Nosych, Andriy A.; Valentino, Gianluca; Aberle, Oliver; Aßmann, Ralph W.; Bertarelli, Alessandro; Boccard, Christian; Bruce, Roderik; Burkart, Florian; Calvo, Eva; Cauchi, Marija; Dallocchio, Alessandro; Deboy, Daniel; Gasior, Marek; Jones, Rhodri; Kain, Verena; Lari, Luisella; Redaelli, Stefano; Rossi, Adriana

    2014-12-01

    At present, the beam-based alignment of the LHC collimators is performed by touching the beam halo with both jaws of each collimator. This method requires dedicated fills at low intensities that are done infrequently and makes this procedure time consuming. This limits the operational flexibility, in particular in the case of changes of optics and orbit configuration in the experimental regions. The performance of the LHC collimation system relies on the machine reproducibility and regular loss maps to validate the settings of the collimator jaws. To overcome these limitations and to allow a continuous monitoring of the beam position at the collimators, a design with jaw-integrated Beam Position Monitors (BPMs) was proposed and successfully tested with a prototype (mock-up) collimator in the CERN SPS. Extensive beam experiments allowed to determine the achievable accuracy of the jaw alignment for single and multi-turn operation. In this paper, the results of these experiments are discussed. The non-linear response of the BPMs is compared to the predictions from electromagnetic simulations. Finally, the measured alignment accuracy is compared to the one achieved with the present collimators in the LHC.

  6. New method for improving angle measurement precision of laser collimation system under complex background

    NASA Astrophysics Data System (ADS)

    Zhao, Xiaofeng; Chen, He; Tan, Lilong; Zhang, Zhili; Cai, Wei

    2014-09-01

    We have proposed a new method for improving angle measurement precision based on the principle of CCD laser collimation in this paper. First, through the control of the laser's state, on or off, by the Digital Signal Processor (DSP), the collimation light and the background light can be sampled, individually. Second, with the comparison between the sampled value of the background light intensity and the threshold value which has been set in the DSP previously, the DSP can automatically control Complex Programmable Logic Device (CPLD) to adjust the light integral time of CCD to adapt to different environment background and the changeable scanning driver of CCD is realized. Last, by the digital wave filtering the impact of the background light on the collimation light can be removed. With the comprehensive application of the controlling technology of automatically changeable scanning driving, collimation light on or off, A/D conversion and adaptive filtering, the integration time of the collimation system can automatically adjust to the proper value according to the change of the environment and the impact of the background light on the collimation system can be well removed. The simulation results show that the new method can achieve the self-adaptable control with the change of the environment and can improve the measurement precision of the laser collimation system under the complex environment.

  7. Detector, collimator and real-time reconstructor for a new scanning-beam digital x-ray (SBDX) prototype

    NASA Astrophysics Data System (ADS)

    Speidel, Michael A.; Tomkowiak, Michael T.; Raval, Amish N.; Dunkerley, David A. P.; Slagowski, Jordan M.; Kahn, Paul; Ku, Jamie; Funk, Tobias

    2015-03-01

    Scanning-beam digital x-ray (SBDX) is an inverse geometry fluoroscopy system for low dose cardiac imaging. The use of a narrow scanned x-ray beam in SBDX reduces detected x-ray scatter and improves dose efficiency, however the tight beam collimation also limits the maximum achievable x-ray fluence. To increase the fluence available for imaging, we have constructed a new SBDX prototype with a wider x-ray beam, larger-area detector, and new real-time image reconstructor. Imaging is performed with a scanning source that generates 40,328 narrow overlapping projections from 71 x 71 focal spot positions for every 1/15 s scan period. A high speed 2-mm thick CdTe photon counting detector was constructed with 320x160 elements and 10.6 cm x 5.3 cm area (full readout every 1.28 μs), providing an 86% increase in area over the previous SBDX prototype. A matching multihole collimator was fabricated from layers of tungsten, brass, and lead, and a multi-GPU reconstructor was assembled to reconstruct the stream of captured detector images into full field-of-view images in real time. Thirty-two tomosynthetic planes spaced by 5 mm plus a multiplane composite image are produced for each scan frame. Noise equivalent quanta on the new SBDX prototype measured 63%-71% higher than the previous prototype. X-ray scatter fraction was 3.9-7.8% when imaging 23.3-32.6 cm acrylic phantoms, versus 2.3- 4.2% with the previous prototype. Coronary angiographic imaging at 15 frame/s was successfully performed on the new SBDX prototype, with live display of either a multiplane composite or single plane image.

  8. Optimum angle-cut of collimator for dense objects in high-energy proton radiography

    NASA Astrophysics Data System (ADS)

    Xu, Hai-Bo; Zheng, Na

    2016-02-01

    The use of minus identity lenses with an angle-cut collimator can achieve high contrast images in high-energy proton radiography. This article presents the principles of choosing the angle-cut aperture of the collimator for different energies and objects. Numerical simulation using the Monte Carlo code Geant4 has been implemented to investigate the entire radiography for the French test object. The optimum angle-cut apertures of the collimators are also obtained for different energies. Supported by NSAF (11176001) and Science and Technology Developing Foundation of China Academy of Engineering Physics (2012A0202006)

  9. Comparison of the TESLA, NLC and CLIC beam-collimation system performance

    SciTech Connect

    Alexandr I Drozhdin; Grahame Blair; Lewis P Keller

    2003-05-28

    This report describes studies performed in the framework of the Collimation Task Force organized to support the work of the International Linear Collider Technical Review Committee. The post-linac beam-collimation systems in the TESLA, JLC/NLC and CLIC linear-collider designs are compared using the same computer code under the same assumptions. Their performance is quantified in terms of beam-halo and synchrotron-radiation collimation efficiency. The performance of the current designs varies across projects, and does not always meet the original design goals. But these comparisons suggest that achieving the required performance in a future linear collider is feasible.

  10. Jets and Photons

    NASA Astrophysics Data System (ADS)

    Ellis, Stephen D.; Roy, Tuhin S.; Scholtz, Jakub

    2013-03-01

    This Letter applies the concept of “jets,” as constructed from calorimeter cell four-vectors, to jets composed (primarily) of photons (or leptons). Thus jets become a superset of both traditional objects such as QCD jets, photons, and electrons, and more unconventional objects such as photon jets and electron jets, defined as collinear photons and electrons, respectively. Since standard objects such as single photons become a subset of jets in this approach, standard jet substructure techniques are incorporated into the photon finder toolbox. Using a (reasonably) realistic calorimeter model we demonstrate that, for a single photon identification efficiency of 80% or above, the use of jet substructure techniques reduces the number of QCD jets faking photons by factors of 2.5 to 4. Depending on the topology of the photon jets, the substructure variables reduce the number of photon jets faking single photons by factors of 10 to 103 at a single photon identification efficiency of 80%.

  11. Magnetically collimated pair jets at the LLNL Titan laser

    NASA Astrophysics Data System (ADS)

    Williams, Jackson; Chen, Hui; Barnak, Daniel; Betti, Riccardo; Fiksel, Gennady; Hazi, Andrew; Kerr, Shaun; Krauland, Christine; Link, Anthony; Manuel, Mario; Meyerhofer, David; Nagel, Sabrina; Park, Jaebum; Peebles, Jonathan; Pollock, Bradley; Tommasini, Riccardo

    2015-11-01

    Positron-electron pair production experiments were performed at the Titan laser at the Jupiter Laser Facility to investigate the dependence of target thickness and atomic number on pair yield. Externally applied axial magnetic fields, generated by a Helmholtz coil, were used to collimate positrons where the signal observed at the detector increased by a factor of 20 over reference shots without a field. This enabled the detection of positrons from a range of target materials. The emitted positron yield was found to be proportional to the square of the atomic number. This scaling is reduced from the Bethe-Heitler cross section of Z4 by Compton scattering and the stopping power of the target. Monte Carlo simulations support these conclusions, providing a power-law scaling of emitted positrons for all materials and a range of mm-thick targets. This work was performed under the auspices of the U.S. Department of Energy (DOE) by Lawrence Livermore National Laboratory under Contract DE-AC52-07NA27344 and funded by the LLNL LDRD program under tracking code 12-ERD-062 and the LLNL LGSP.

  12. A collimated neutron detector for RFP plasmas in MST

    NASA Astrophysics Data System (ADS)

    Capecchi, W. J.; Anderson, J. K.; Bonofiglo, P. J.; Kim, J.; Sears, S.

    2016-11-01

    The neutron emissivity profile in the Madison Symmetric Torus is being reconstructed through the use of a collimated neutron detector. A scintillator-photomultiplier tube (PMT) system is employed to detect the fusion neutrons with the plasma viewing volume defined by a 55 cm deep, 5 cm diameter aperture. Effective detection of neutrons from the viewing volume is achieved through neutron moderation using 1300 lbs of high density polyethylene shielding, which modeling predicts attenuates the penetrating flux by a factor of 104 or more. A broad spectrum of gamma radiation is also present due to the unconfined fusion proton bombardment of the thick aluminum vacuum vessel. A 15 cm cylindrical liquid scintillator of 3.8 cm diameter is used to further increase directional sensitivity. A fast (5 ns rise time) preamplifier and digitization at 500 MHz prevent pulse pile-up even at high count rates (˜104/s). The entire neutron camera system is situated on an adjustable inclining base which provides the differing plasma viewing volumes necessary for reconstruction of the neutron emissivity profile. This profile, directly related to the fast-ion population, allows for an investigation of the critical fast-ion pressure gradient required to destabilize a neutral beam driven Alfvénic mode which has been shown to transport fast ions.

  13. Novel adaptive fiber-optics collimator for coherent beam combination.

    PubMed

    Zhi, Dong; Ma, Pengfei; Ma, Yanxing; Wang, Xiaolin; Zhou, Pu; Si, Lei

    2014-12-15

    In this manuscript, we experimentally validate a novel design of adaptive fiber-optics collimator (AFOC), which utilizes two levers to enlarge the movable range of the fiber end cap. The enlargement of the range makes the new AFOC possible to compensate the end-cap/tilt aberration in fiber laser beam combining system. The new AFOC based on flexible hinges and levers was fabricated and the performance of the new AFOC was tested carefully, including its control range, frequency response and control accuracy. Coherent beam combination (CBC) of two 5-W fiber amplifiers array with simultaneously end-cap/tilt control and phase-locking control was implemented successfully with the novel AFOC. Experimental results show that the average normalized power in the bucket (PIB) value increases from 0.311 to 0.934 with active phasing and tilt aberration compensation simultaneously, and with both controls on, the fringe contrast improves to more than 82% from 0% for the case with both control off. This work presents a promising structure for tilt aberration control in high power CBC system.

  14. RADBALLTECHNOLOGY TESTING AND MCNP MODELING OF THE TUNGSTEN COLLIMATOR

    SciTech Connect

    Farfan, E.

    2010-07-08

    The United Kingdom's National Nuclear Laboratory (NNL) has developed a remote, non-electrical, radiation-mapping device known as RadBall{trademark}, which can locate and quantify radioactive hazards within contaminated areas of the nuclear industry. RadBall{trademark} consists of a colander-like outer shell that houses a radiation-sensitive polymer sphere. The outer shell works to collimate radiation sources and those areas of the polymer sphere that are exposed react, becoming increasingly more opaque, in proportion to the absorbed dose. The polymer sphere is imaged in an optical-CT scanner, which produces a high resolution 3D map of optical attenuation coefficients. Subsequent analysis of the optical attenuation matrix provides information on the spatial distribution of sources in a given area forming a 3D characterization of the area of interest. RadBall{trademark} has no power requirements and can be positioned in tight or hard-to reach locations. The RadBall{trademark} technology has been deployed in a number of technology trials in nuclear waste reprocessing plants at Sellafield in the United Kingdom and facilities of the Savannah River National Laboratory (SRNL). This study focuses on the RadBall{trademark} testing and modeling accomplished at SRNL.

  15. Collimating cylindrical diffractive lenses: rigorous electromagnetic analysis and scalar approximation.

    PubMed

    Glytsis, E N; Harrigan, M E; Hirayama, K; Gaylord, T K

    1998-01-01

    Practical collimating diffractive cylindrical lenses of 2, 4, 8, and 16 discrete levels are analyzed with a sequential application of the two-region formulation of the rigorous electromagnetic boundary-element method (BEM). A Gaussian beam of TE or TM polarization is incident upon the finite-thickness lens. F/4, F/2, and F/1.4 lenses are analyzed and near-field electric-field patterns are presented. The near-field wave-front quality is quantified by its mean-square deviation from a planar wave front. This deviation is found to be less than 0.05 free-space wavelengths. The far-field intensity patterns are determined and compared with the ones predicted by the approximate Fraunhofer scalar diffraction analysis. The diffraction efficiencies determined with the rigorous BEM are found to be generally lower than those obtained with the scalar approximation. For comparison, the performance characteristics of the corresponding continuous Fresnel (continuous profile within a zone but discontinuous at zone boundaries) and continuous refractive lenses are determined by the use of both the BEM and the scalar approximation. The diffraction efficiency of the continuous Fresnel lens is found to be similar to that of the 16-level diffractive lens but less than that of the continuous refractive lens. It is shown that the validity of the scalar approximation deteriorates as the lens f-number decreases.

  16. Simulation study comparing the imaging performance of a solid state detector with a rotating slat collimator versus parallel beam collimator setups

    NASA Astrophysics Data System (ADS)

    Staelens, Steven; Vandenberghe, Stefaan; De Beenhouwer, Jan; De Clercq, Stijn; D'Asseler, Yves; Lemahieu, Ignace; Van de Walle, Rik

    2004-05-01

    The main goal of this work is to assess the overall imaging performance of dedicated new solid state devices compared to a traditional scintillation camera for use in SPECT imaging. A solid state detector with a rotating slat collimator will be compared with the same detector mounted with a classical collimator as opposed to a traditional Anger camera. A better energy resolution characterizes the solid state materials while the rotating slat collimator promises a better sensitivity-resolution tradeoff. The evaluation of the different imaging modalities is done using GATE, a recently developed Monte Carlo code. Several features for imaging performance evaluation were addressed: spatial resolution, energy resolution, sensitivity, and a ROC analysis was performed to evaluate the hot spot detectability. In this way a difference in perfromance was concluded for the diverse imaging techniques which allows a task dependent application of these modalities in future clinical practice.

  17. Resonance formation in photon-photon collisions

    SciTech Connect

    Gidal, G.

    1988-08-01

    Recent experimental progress on resonance formation in photon-photon collisions is reviewed with particular emphasis on the pseudoscalar and tensor nonents and on the ..gamma gamma..* production of spin-one resonances. 37 refs., 17 figs., 5 tabs.

  18. Collimation testing by use of the Lau effect coupled with moire readout

    SciTech Connect

    Rana, Santosh; Prakash, Shashi

    2006-05-10

    We present an easy, simple, and inexpensive technique for checking the quality of the collimation of optical beams using the Lau effect combined with moire readout. The experimental arrangement consists of a modified Lau-based interferometer in which a white-light incoherent source illuminates a set of two gratings. A collimating lens is placed between the two gratings such that the self-images of the second grating are formed. The third grating is positioned at one of the self-imaging planes forming moire fringes. The type of the moire fringe demonstrates the quality of collimation of the optical beam. The necessary theoretical background is presented and the results of our experimental investigation are reported. The technique can also be used for accurate determination of the focal length of a collimating lens using low-cost components.

  19. Successive approximation algorithm for beam-position-monitor-based LHC collimator alignment

    NASA Astrophysics Data System (ADS)

    Valentino, Gianluca; Nosych, Andriy A.; Bruce, Roderik; Gasior, Marek; Mirarchi, Daniele; Redaelli, Stefano; Salvachua, Belen; Wollmann, Daniel

    2014-02-01

    Collimators with embedded beam position monitor (BPM) button electrodes will be installed in the Large Hadron Collider (LHC) during the current long shutdown period. For the subsequent operation, BPMs will allow the collimator jaws to be kept centered around the beam orbit. In this manner, a better beam cleaning efficiency and machine protection can be provided at unprecedented higher beam energies and intensities. A collimator alignment algorithm is proposed to center the jaws automatically around the beam. The algorithm is based on successive approximation and takes into account a correction of the nonlinear BPM sensitivity to beam displacement and an asymmetry of the electronic channels processing the BPM electrode signals. A software implementation was tested with a prototype collimator in the Super Proton Synchrotron. This paper presents results of the tests along with some considerations for eventual operation in the LHC.

  20. Improving the performance of gold nanohole array biosensors by controlling the optical collimation conditions.

    PubMed

    Weber de Menezes, Jacson; Thesing, Anderson; Valsecchi, Chiara; Armas, Luis E G; Brolo, Alexandre G

    2015-07-20

    An experimental investigation on how the bulk and surface sensitivities of gold nanohole arrays fabricated by interference lithography affect the degree of white light beam collimation is presented. The optical transmission response of nanohole arrays has been recorded by focused and collimated beam transmission spectra. The results show that both the bulk and surface sensitivities for the collimated case are much larger than for the focused case. In particular, the shape of the spectra was dependent on the degree of beam collimation. The results showed that improved sensing performance (around 3.5 times) and higher figure of merit (around 4.4 times) can be obtained by simply adjusting the incident/collection experimental conditions in transmission measurements. PMID:26367835

  1. Optimization of the collimation system for CSNS/RCS with the robust conjugate direction search algorithm

    NASA Astrophysics Data System (ADS)

    Ji, Hong-Fei; Jiao, Yi; Huang, Ming-Yang; Xu, Shou-Yan; Wang, Na; Wang, Sheng

    2016-09-01

    The Robust Conjugate Direction Search (RCDS) method is used to optimize the collimation system for the Rapid Cycling Synchrotron (RCS) of the China Spallation Neutron Source (CSNS). The parameters of secondary collimators are optimized for a better performance of the collimation system. To improve the efficiency of the optimization, the Objective Ring Beam Injection and Tracking (ORBIT) parallel module combined with MATLAB parallel computing is used, which can run multiple ORBIT instances simultaneously. This study presents a way to find an optimal parameter combination of the secondary collimators for a machine model in preparation for CSNS/RCS commissioning. Supported by National Natural Science Foundation of China (11475202, 11405187, 11205185) and Youth Innovation Promotion Association of Chinese Academy of Sciences (2015009)

  2. Improving the performance of gold nanohole array biosensors by controlling the optical collimation conditions.

    PubMed

    Weber de Menezes, Jacson; Thesing, Anderson; Valsecchi, Chiara; Armas, Luis E G; Brolo, Alexandre G

    2015-07-20

    An experimental investigation on how the bulk and surface sensitivities of gold nanohole arrays fabricated by interference lithography affect the degree of white light beam collimation is presented. The optical transmission response of nanohole arrays has been recorded by focused and collimated beam transmission spectra. The results show that both the bulk and surface sensitivities for the collimated case are much larger than for the focused case. In particular, the shape of the spectra was dependent on the degree of beam collimation. The results showed that improved sensing performance (around 3.5 times) and higher figure of merit (around 4.4 times) can be obtained by simply adjusting the incident/collection experimental conditions in transmission measurements.

  3. Distance and depth measurements of radioactive objects utilizing rotating slant-hole collimators.

    PubMed

    Lancaster, J L; Kopp, D T; Lasher, J C; Blumhardt, R

    1988-01-01

    It is desirable to have an accurate method to determine the depth of various organs which are commonly imaged in nuclear medicine. The method presented here utilizes the geometrical characteristics of the rotating slant-hole collimator to calculate distances and depth. The accuracy of the calculated distance was evaluated for point sources in air separated by 11 cm. The reproducibility of the calculated distance was evaluated as a function of the extent of collimator rotation between observations for angular differences of 45 degrees, 90 degrees, 135 degrees, and 180 degrees. The theory of the method is discussed and the relative error in distance calculation analyzed mathematically as a function of (1) collimator rotation angle error, (2) collimator slant angle error, and (3) position calculation error. Our findings indicate this method to be accurate to within 2% with a reproducibility range of 3%-5% for point sources.

  4. Collimation system design for beam loss localization with slipstacking injection in the Fermilab Main Injector

    SciTech Connect

    Drozhdin, A.I.; Brown, B.C.; Johnson, D.E.; Koba, K.; Kourbanis, I.; Mokhov, N.V.; Rakhno, I.L.; Sidorov, V.I.; /Fermilab

    2007-06-01

    Results of modeling with the 3-D STRUCT and MARS15 codes of beam loss localization and related radiation effects are presented for the slipstacking injection to the Fermilab Main Injector. Simulations of proton beam loss are done using multi-turn tracking with realistic accelerator apertures, nonlinear fields in the accelerator magnets and time function of the RF manipulations to explain the results of beam loss measurements. The collimation system consists of one primary and four secondary collimators. It intercepts a beam power of 1.6 kW at a scraping rate of 5% of 5.5E+13 ppp, with a beam loss rate in the ring outside the collimation region of 1 W/m or less. Based on thorough energy deposition and radiation modeling, a corresponding collimator design was developed that satisfies all the radiation and engineering constraints.

  5. Collimator design for spatially-fractionated proton beams for radiobiology research

    NASA Astrophysics Data System (ADS)

    Lee, Eunsin; Meyer, Juergen; Sandison, George

    2016-07-01

    Preclinical and translational research is an imperative to improve the efficacy of proton radiotherapy. We present a feasible and practical method to produce spatially-modulated proton beams for cellular and small animal research for clinical and research facilities. The University of Washington (UW) 50.5 MeV proton research beamline hosting a brass collimation system was modeled using Monte Carlo simulations. This collimator consisted of an array of 2 cm long slits to cover an area of 2  ×  2 cm2. To evaluate the collimator design effects on dose rate, valley dose and the peak-to-valley dose ratios (PVDR) the following parameters were varied; slit width (0.1–1.0 mm), peak center-to-center distance (1–3 mm), collimator thickness (1–7 cm) and collimator location along the beam axis. Several combinations of slit widths and 1 mm spacing achieved uniform dose at the Bragg peak while maintaining spatial modulation on the beam entrance. A more detailed analysis was carried out for the case of a slit width of 0.3 mm, peak center-to-center distance of 1 mm, a collimator thickness of 5 cm and with the collimator flush against the water phantom. The dose rate at 5 mm depth dropped relative to an open field by a factor of 12 and produced a PVDR of 10.1. Technical realization of proton mini-beams for radiobiology small animal research is demonstrated to be feasible. It is possible to obtain uniform dose at depth while maintaining reasonable modulation at shallower depths near the beam entrance. While collimator design is important the collimator location has a strong influence on the entrance region PVDRs and on dose rate. These findings are being used to manufacture a collimator for installation on the UW cyclotron proton beam nozzle. This collimator will enable comparative studies on the radiobiological efficacy of x-rays and proton beams.

  6. Collimator design for spatially-fractionated proton beams for radiobiology research.

    PubMed

    Lee, Eunsin; Meyer, Juergen; Sandison, George

    2016-07-21

    Preclinical and translational research is an imperative to improve the efficacy of proton radiotherapy. We present a feasible and practical method to produce spatially-modulated proton beams for cellular and small animal research for clinical and research facilities. The University of Washington (UW) 50.5 MeV proton research beamline hosting a brass collimation system was modeled using Monte Carlo simulations. This collimator consisted of an array of 2 cm long slits to cover an area of 2  ×  2 cm(2). To evaluate the collimator design effects on dose rate, valley dose and the peak-to-valley dose ratios (PVDR) the following parameters were varied; slit width (0.1-1.0 mm), peak center-to-center distance (1-3 mm), collimator thickness (1-7 cm) and collimator location along the beam axis. Several combinations of slit widths and 1 mm spacing achieved uniform dose at the Bragg peak while maintaining spatial modulation on the beam entrance. A more detailed analysis was carried out for the case of a slit width of 0.3 mm, peak center-to-center distance of 1 mm, a collimator thickness of 5 cm and with the collimator flush against the water phantom. The dose rate at 5 mm depth dropped relative to an open field by a factor of 12 and produced a PVDR of 10.1. Technical realization of proton mini-beams for radiobiology small animal research is demonstrated to be feasible. It is possible to obtain uniform dose at depth while maintaining reasonable modulation at shallower depths near the beam entrance. While collimator design is important the collimator location has a strong influence on the entrance region PVDRs and on dose rate. These findings are being used to manufacture a collimator for installation on the UW cyclotron proton beam nozzle. This collimator will enable comparative studies on the radiobiological efficacy of x-rays and proton beams. PMID:27362834

  7. Collimator design for spatially-fractionated proton beams for radiobiology research

    NASA Astrophysics Data System (ADS)

    Lee, Eunsin; Meyer, Juergen; Sandison, George

    2016-07-01

    Preclinical and translational research is an imperative to improve the efficacy of proton radiotherapy. We present a feasible and practical method to produce spatially-modulated proton beams for cellular and small animal research for clinical and research facilities. The University of Washington (UW) 50.5 MeV proton research beamline hosting a brass collimation system was modeled using Monte Carlo simulations. This collimator consisted of an array of 2 cm long slits to cover an area of 2  ×  2 cm2. To evaluate the collimator design effects on dose rate, valley dose and the peak-to-valley dose ratios (PVDR) the following parameters were varied; slit width (0.1-1.0 mm), peak center-to-center distance (1-3 mm), collimator thickness (1-7 cm) and collimator location along the beam axis. Several combinations of slit widths and 1 mm spacing achieved uniform dose at the Bragg peak while maintaining spatial modulation on the beam entrance. A more detailed analysis was carried out for the case of a slit width of 0.3 mm, peak center-to-center distance of 1 mm, a collimator thickness of 5 cm and with the collimator flush against the water phantom. The dose rate at 5 mm depth dropped relative to an open field by a factor of 12 and produced a PVDR of 10.1. Technical realization of proton mini-beams for radiobiology small animal research is demonstrated to be feasible. It is possible to obtain uniform dose at depth while maintaining reasonable modulation at shallower depths near the beam entrance. While collimator design is important the collimator location has a strong influence on the entrance region PVDRs and on dose rate. These findings are being used to manufacture a collimator for installation on the UW cyclotron proton beam nozzle. This collimator will enable comparative studies on the radiobiological efficacy of x-rays and proton beams.

  8. Simulation, design, and testing of a high power collimator for the RDS-112 cyclotron.

    PubMed

    Peeples, Johanna L; Stokely, Matthew H; Poorman, Michael C; Bida, Gerald T; Wieland, Bruce W

    2015-03-01

    A high power [F-18] fluoride target package for the RDS-112 cyclotron has been designed, tested, and commercially deployed. The upgrade includes the CF-1000 target, a 1.3kW water target with an established commercial history on RDS-111/Eclipse cyclotrons, and a redesigned collimator with improved heat rejection capabilities. Conjugate heat transfer analyses were employed to both evaluate the existing collimator capabilities and design a suitable high current replacement.

  9. Simulation of the ILC Collimation System using BDSIM, MARS15 and STRUCT

    SciTech Connect

    Carter, J.; Agapov, I.; Blair, G.A.; Deacon, L.; Drozhdin, A.I.; Mokhov, N.V.; Nosochkov, Y.M.; Seryi, A.A.; /SLAC

    2006-07-12

    The simulation codes BDSIM, MARS15 and STRUCT are used to simulate in detail the collimation section of the International Linear Collider (ILC). A comparative study of the collimation system performance for the 250 x 250 GeV machine is conducted, and the key radiation loads are calculated. Results for the latest ILC designs are presented together with their implications for future design iterations.

  10. Laser beam collimation using a phase conjugate Twyman-Green interferometer

    NASA Technical Reports Server (NTRS)

    Shukla, R. P.; Dokhanian, M.; George, M. C.; Venkateswarlu, Putcha

    1991-01-01

    This paper presents an improved technique for testing laser beam collimation using a phase conjugate Twyman-Green interferometer. The technique is useful for measuring laser beam divergence. It is possible using this technique to detect the defocusing of the order of one micrometer for a well corrected collimating lens. A relation is derived for the defocusing that can be detected by the phase conjugate interferometer.

  11. Impact of the A48 collimator on the Tevatron B0 dipoles

    SciTech Connect

    Ludovic Y. Nicolas; Nikolai V. Mokhov

    2003-06-23

    To protect the CDF detector components in an event of an abort kicker prefire (AKP) in the Tevatron, a new collimator is to be installed at the A48 location during the summer 2003 shutdown. Detailed calculations have shown that this 0.5-m long ''single L-shape'' steel collimator will intercept a bunch of protons when such an incident occurs, providing reliable protection of the CDF main detector at an AKP. It will also mitigate the backgrounds induced by elastic beam-gas interactions upstream of B0. Although the Roman Pot detectors downstream of the A48 collimator will see an increased background, the amount of radiation they will receive either resulting from beam halo interactions in the collimator or during an AKP will not damage their sensitive parts. Secondaries resulting from beam halo interactions with the A48 collimator do not noticeably affect the downstream dipoles. The case of an AKP is quite different. As opposed to halo hits in the ''single-L shape'' unit (around 10{sup 5} p/s), a bunch lost on A48 during an AKP represents more than 2 x 10{sup 11} protons ''instantaneously'' interacting with the collimator material. Although the collimator protects the downstream superconducting (SC) dipoles against a damage in such an event, secondaries generated in A48 create a significant radiation load on the dipoles which will most likely result in a quench of the first one. This effect is studied in detail in this note. Energy deposition in the B0 dipoles downstream of a new A48 collimator to be installed in the Tevatron this summer is calculated with the MARS14 code to evaluate the dipole's quench stability at an abort kicker prefire.

  12. Study of High-Frequency Impedance of Small-Angle Tapers and Collimators

    SciTech Connect

    Stupakov, Gennady; Podobedov, B.; /Brookhaven

    2010-06-04

    Collimators and transitions in accelerator vacuum chambers often include small-angle tapering to lower the wakefields generated by the beam. While the low-frequency impedance is well described by Yokoya's formula (for axisymmetric geometry), much less is known about the behavior of the impedance in the high frequency limit. In this paper we develop an analytical approach to the highfrequency regime for round collimators and tapers. Our analytical results are compared with computer simulations using the code ECHO.

  13. Simulation, Design, and Testing of a High Power Collimator for the RDS-112 Cyclotron

    PubMed Central

    Peeples, Johanna L.; Stokely, Matthew H.; Poorman, Michael C.; Bida, Gerald T.; Wieland, Bruce W.

    2015-01-01

    A high power [F-18]fluoride target package for the RDS-112 cyclotron has been designed, tested, and commercially deployed. The upgrade includes the CF-1000 target, a 1.3 kW water target with an established commercial history on RDS-111/Eclipse cyclotrons, and a redesigned collimator with improved heat rejection capabilities. Conjugate heat transfer analyses were employed to both evaluate the existing collimator capabilities and design a suitable high current replacement. PMID:25562677

  14. Modules and methods for all photonic computing

    DOEpatents

    Schultz, David R.; Ma, Chao Hung

    2001-01-01

    A method for all photonic computing, comprising the steps of: encoding a first optical/electro-optical element with a two dimensional mathematical function representing input data; illuminating the first optical/electro-optical element with a collimated beam of light; illuminating a second optical/electro-optical element with light from the first optical/electro-optical element, the second optical/electro-optical element having a characteristic response corresponding to an iterative algorithm useful for solving a partial differential equation; iteratively recirculating the signal through the second optical/electro-optical element with light from the second optical/electro-optical element for a predetermined number of iterations; and, after the predetermined number of iterations, optically and/or electro-optically collecting output data representing an iterative optical solution from the second optical/electro-optical element.

  15. Thermomechanical response of Large Hadron Collider collimators to proton and ion beam impacts

    NASA Astrophysics Data System (ADS)

    Cauchi, Marija; Assmann, R. W.; Bertarelli, A.; Carra, F.; Cerutti, F.; Lari, L.; Redaelli, S.; Mollicone, P.; Sammut, N.

    2015-04-01

    The CERN Large Hadron Collider (LHC) is designed to accelerate and bring into collision high-energy protons as well as heavy ions. Accidents involving direct beam impacts on collimators can happen in both cases. The LHC collimation system is designed to handle the demanding requirements of high-intensity proton beams. Although proton beams have 100 times higher beam power than the nominal LHC lead ion beams, specific problems might arise in case of ion losses due to different particle-collimator interaction mechanisms when compared to protons. This paper investigates and compares direct ion and proton beam impacts on collimators, in particular tertiary collimators (TCTs), made of the tungsten heavy alloy INERMET® 180. Recent measurements of the mechanical behavior of this alloy under static and dynamic loading conditions at different temperatures have been done and used for realistic estimates of the collimator response to beam impact. Using these new measurements, a numerical finite element method (FEM) approach is presented in this paper. Sequential fast-transient thermostructural analyses are performed in the elastic-plastic domain in order to evaluate and compare the thermomechanical response of TCTs in case of critical beam load cases involving proton and heavy ion beam impacts.

  16. An empirical model for calculation of the collimator contamination dose in therapeutic proton beams.

    PubMed

    Vidal, M; De Marzi, L; Szymanowski, H; Guinement, L; Nauraye, C; Hierso, E; Freud, N; Ferrand, R; François, P; Sarrut, D

    2016-02-21

    Collimators are used as lateral beam shaping devices in proton therapy with passive scattering beam lines. The dose contamination due to collimator scattering can be as high as 10% of the maximum dose and influences calculation of the output factor or monitor units (MU). To date, commercial treatment planning systems generally use a zero-thickness collimator approximation ignoring edge scattering in the aperture collimator and few analytical models have been proposed to take scattering effects into account, mainly limited to the inner collimator face component. The aim of this study was to characterize and model aperture contamination by means of a fast and accurate analytical model. The entrance face collimator scatter distribution was modeled as a 3D secondary dose source. Predicted dose contaminations were compared to measurements and Monte Carlo simulations. Measurements were performed on two different proton beam lines (a fixed horizontal beam line and a gantry beam line) with divergent apertures and for several field sizes and energies. Discrepancies between analytical algorithm dose prediction and measurements were decreased from 10% to 2% using the proposed model. Gamma-index (2%/1 mm) was respected for more than 90% of pixels. The proposed analytical algorithm increases the accuracy of analytical dose calculations with reasonable computation times.

  17. An empirical model for calculation of the collimator contamination dose in therapeutic proton beams

    NASA Astrophysics Data System (ADS)

    Vidal, M.; De Marzi, L.; Szymanowski, H.; Guinement, L.; Nauraye, C.; Hierso, E.; Freud, N.; Ferrand, R.; François, P.; Sarrut, D.

    2016-02-01

    Collimators are used as lateral beam shaping devices in proton therapy with passive scattering beam lines. The dose contamination due to collimator scattering can be as high as 10% of the maximum dose and influences calculation of the output factor or monitor units (MU). To date, commercial treatment planning systems generally use a zero-thickness collimator approximation ignoring edge scattering in the aperture collimator and few analytical models have been proposed to take scattering effects into account, mainly limited to the inner collimator face component. The aim of this study was to characterize and model aperture contamination by means of a fast and accurate analytical model. The entrance face collimator scatter distribution was modeled as a 3D secondary dose source. Predicted dose contaminations were compared to measurements and Monte Carlo simulations. Measurements were performed on two different proton beam lines (a fixed horizontal beam line and a gantry beam line) with divergent apertures and for several field sizes and energies. Discrepancies between analytical algorithm dose prediction and measurements were decreased from 10% to 2% using the proposed model. Gamma-index (2%/1 mm) was respected for more than 90% of pixels. The proposed analytical algorithm increases the accuracy of analytical dose calculations with reasonable computation times.

  18. Experimental study of phase locking of fiber collimators using internal beam-tail interference

    NASA Astrophysics Data System (ADS)

    Beresnev, Leonid A.; Vorontsov, Mikhail A.; Weyrauch, Thomas; Carhart, Gary; Lachinova, Svetlana L.; Liu, Jiang

    2011-02-01

    Experimental study of internal phase locking of a seven fiber collimator array is presented. As a metric for the feedback loop the periphery areas (tails) of beams outgoing from three adjacent fiber tips are used before the beams are clipped by the lens apertures. The "intercepted" tails of beams are redirected back into the collimator array forming an interference pattern located between adjacent collimators. Optical energy from one region of the pattern is selected by a pinhole, detected with a photo-diode and used as a metric signal for an SPGD controller to lock the phase of the three adjacent beam tails. The non-common phase difference of the outgoing wavefronts from these three collimators can be manipulated by altering the position of the pinhole in focal plane of the interference pattern and is removed (set to "zero" or 2π increments) by a displacement selected to produce the expected far field interference pattern. To phase lock the beams from seven collimators arranged in a hexagonal array, three pinholes, each of which receives some light from the center collimator are used. A sum of the signals from the three photo-diodes placed behind these pinholes is used to lock the phase of the six periphery beams to the central uncontrolled reference without beam splitters or a remote target-in-theloop metric.

  19. Efficient dielectric metasurface collimating lenses for mid-infrared quantum cascade lasers.

    PubMed

    Arbabi, Amir; Briggs, Ryan M; Horie, Yu; Bagheri, Mahmood; Faraon, Andrei

    2015-12-28

    Light emitted from single-mode semiconductor lasers generally has large divergence angles, and high numerical aperture lenses are required for beam collimation. Visible and near infrared lasers are collimated using aspheric glass or plastic lenses, yet collimation of mid-infrared quantum cascade lasers typically requires more costly aspheric lenses made of germanium, chalcogenide compounds, or other infrared-transparent materials. Here we report mid-infrared dielectric metasurface flat lenses that efficiently collimate the output beam of single-mode quantum cascade lasers. The metasurface lenses are composed of amorphous silicon posts on a flat sapphire substrate and can be fabricated at low cost using a single step conventional UV binary lithography. Mid-infrared radiation from a 4.8 μm distributed-feedback quantum cascade laser is collimated using a polarization insensitive metasurface lens with 0.86 numerical aperture and 79% transmission efficiency. The collimated beam has a half divergence angle of 0.36° and beam quality factor of M2=1.02.

  20. Efficient dielectric metasurface collimating lenses for mid-infrared quantum cascade lasers.

    PubMed

    Arbabi, Amir; Briggs, Ryan M; Horie, Yu; Bagheri, Mahmood; Faraon, Andrei

    2015-12-28

    Light emitted from single-mode semiconductor lasers generally has large divergence angles, and high numerical aperture lenses are required for beam collimation. Visible and near infrared lasers are collimated using aspheric glass or plastic lenses, yet collimation of mid-infrared quantum cascade lasers typically requires more costly aspheric lenses made of germanium, chalcogenide compounds, or other infrared-transparent materials. Here we report mid-infrared dielectric metasurface flat lenses that efficiently collimate the output beam of single-mode quantum cascade lasers. The metasurface lenses are composed of amorphous silicon posts on a flat sapphire substrate and can be fabricated at low cost using a single step conventional UV binary lithography. Mid-infrared radiation from a 4.8 μm distributed-feedback quantum cascade laser is collimated using a polarization insensitive metasurface lens with 0.86 numerical aperture and 79% transmission efficiency. The collimated beam has a half divergence angle of 0.36° and beam quality factor of M2=1.02. PMID:26831996

  1. An empirical model for calculation of the collimator contamination dose in therapeutic proton beams.

    PubMed

    Vidal, M; De Marzi, L; Szymanowski, H; Guinement, L; Nauraye, C; Hierso, E; Freud, N; Ferrand, R; François, P; Sarrut, D

    2016-02-21

    Collimators are used as lateral beam shaping devices in proton therapy with passive scattering beam lines. The dose contamination due to collimator scattering can be as high as 10% of the maximum dose and influences calculation of the output factor or monitor units (MU). To date, commercial treatment planning systems generally use a zero-thickness collimator approximation ignoring edge scattering in the aperture collimator and few analytical models have been proposed to take scattering effects into account, mainly limited to the inner collimator face component. The aim of this study was to characterize and model aperture contamination by means of a fast and accurate analytical model. The entrance face collimator scatter distribution was modeled as a 3D secondary dose source. Predicted dose contaminations were compared to measurements and Monte Carlo simulations. Measurements were performed on two different proton beam lines (a fixed horizontal beam line and a gantry beam line) with divergent apertures and for several field sizes and energies. Discrepancies between analytical algorithm dose prediction and measurements were decreased from 10% to 2% using the proposed model. Gamma-index (2%/1 mm) was respected for more than 90% of pixels. The proposed analytical algorithm increases the accuracy of analytical dose calculations with reasonable computation times. PMID:26816191

  2. Channeling and Volume Reflection Based Crystal Collimation of Tevatron Circulating Beam Halo (T-980)

    SciTech Connect

    Shiltsev, V.; Annala, G.; Drozhdin, A.; Johnson, T.; Legan, A.; Mokhov, N.; Reilly, R.; Still, D.; Tesarek, R.; Zagel, J.; Peggs, S.; /Brookhaven /CERN /Serpukhov, IHEP /INFN, Ferrara /PNPI, CSTD

    2010-05-01

    The T980 crystal collimation experiment is underway at the Tevatron to determine if this technique could increase 980 GeV beam-halo collimation efficiency at high-energy hadron colliders such as the Tevatron and the LHC. T980 also studies various crystal types and parameters. The setup has been substantially enhanced during the Summer 2009 shutdown by installing a new O-shaped crystal in the horizontal goniometer, as well as adding a vertical goniometer with two alternating crystals (O-shaped and multi-strip) and additional beam diagnostics. First measurements with the new system are quite encouraging, with channeled and volume-reflected beams observed on the secondary collimators as predicted. Investigation of crystal collimation efficiencies with crystals in volume reflection and channeling modes are described in comparison with an amorphous primary collimator. Results on the system performance are presented for the end-of-store studies and for entire collider stores. The first investigation of colliding beam collimation simultaneously using crystals in both the vertical and horizontal plane has been made in the regime with horizontally channeled and vertically volume-reflected beams. Planning is underway for significant hardware improvements during the FY10 summer shutdown and for dedicated studies during the final year of Tevatron operation and also for a 'post-collider beam physics running' period.

  3. Single- and double-slit collimating effects on fast-atom diffraction spectra

    NASA Astrophysics Data System (ADS)

    Gravielle, M. S.; Miraglia, J. E.

    2016-09-01

    Diffraction patterns produced by fast He atoms grazingly impinging on a LiF(0 0 1) surface are investigated focusing on the influence of the beam collimation. Single- and double-slit collimating devices situated in front of the beam source are considered. To describe the scattering process we use the Surface Initial Value Representation (SIVR) approximation, which is a semi-quantum approach that incorporates a realistic description of the initial wave packet in terms of the collimating parameters. Our initial wave-packet model is based on the Van Cittert-Zernike theorem. For a single-slit collimation the width of the collimating aperture controls the shape of the azimuthal angle distribution, making different interference mechanisms visible, while the length of the slit affects the polar angle distribution. Additionally, we found that by means of a double-slit collimation it might be possible to obtain a wide polar angle distribution, which is associated with a large spread of the initial momentum perpendicular to the surface, derived from the uncertainty principle. It might be used as a simple way to probe the surface potential for different normal energies.

  4. Photon-jet ridge at RHIC and the LHC

    NASA Astrophysics Data System (ADS)

    Rezaeian, Amir H.

    2016-05-01

    We investigate long-range rapidity correlations of pairs of prompt photons and jets in the color glass condensate framework in proton-proton and proton-nucleus collisions at RHIC and the LHC. We show that photon-jet correlations exhibit long-range azimuthal collimation at the near side for low transverse momenta of the produced photon and jet in high-multiplicity events. These ridge-like features are strikingly similar to the observed ridge effect for dihadron correlations at RHIC and the LHC. We show that correlations in the relative rapidity and the relative azimuthal angle between pairs of prompt photons and jets strongly depend on the gluon saturation dynamics at small-x kinematics and such measurements can help to understand the true origin of the observed dihadron ridge in p +A collisions, and address whether the ridge is a universal phenomenon for all two-particle correlations at high-energy and high-multiplicity events. We also investigate whether there is a ridge-like structure for photon-hadron pair correlations at RHIC and the LHC. We find that the hadronization of jets has nontrivial effects on the photon-jet correlations.

  5. Gantry and isocenter displacements of a linear accelerator caused by an add-on micromultileaf collimator

    SciTech Connect

    Riis, Hans L.; Zimmermann, Sune J.; Hjelm-Hansen, Mogens

    2013-03-15

    Purpose: The delivery of high quality stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT) treatments to the patient requires knowledge of the position of the isocenter to submillimeter accuracy. To meet the requirements the deviation between the radiation and mechanical isocenters must be less than 1 mm. The use of add-on micromultileaf collimators ({mu}MLCs) in SRS and SRT is an additional challenge to the anticipated high-level geometric and dosimetric accuracy of the treatment. The aim of this work was to quantify the gantry excursions during rotation with and without an add-on {mu}MLC attached to the gantry head. In addition, the shift in the position of the isocenter and its correlation to the kV beam center of the cone-beam CT system was included in the study. Methods: The quantification of the gantry rotational performance was done using a pointer supported by an in-house made rigid holder attached to the gantry head of the accelerator. The pointer positions were measured using a digital theodolite. To quantify the effect of an {mu}MLC of 50 kg, the measurements were repeated with the {mu}MLC attached to the gantry head. The displacement of the isocenter due to an add-on {mu}MLC of 50 kg was also investigated. In case of the pointer measurement the {mu}MLC was simulated by weights attached to the gantry head. A method of least squares was applied to determine the position and displacement of the mechanical isocenter. Additionally, the displacement of the radiation isocenter was measured using a ball-bearing phantom and the electronic portal image device system. These measurements were based on 8 MV photon beams irradiated onto the ball from the four cardinal angles and two opposed collimator angles. The measurements and analysis of the data were carried out automatically using software delivered by the manufacturer. Results: The displacement of the mechanical isocenter caused by a 50 kg heavy {mu}MLC was found to be (-0.01 {+-} 0.05, -0

  6. Mini-beam collimator enables microcrystallography experiments on standard beamlines

    PubMed Central

    Fischetti, Robert F.; Xu, Shenglan; Yoder, Derek W.; Becker, Michael; Nagarajan, Venugopalan; Sanishvili, Ruslan; Hilgart, Mark C.; Stepanov, Sergey; Makarov, Oleg; Smith, Janet L.

    2009-01-01

    The high-brilliance X-ray beams from undulator sources at third-generation synchrotron facilities are excellent tools for solving crystal structures of important and challenging biological macromolecules and complexes. However, many of the most important structural targets yield crystals that are too small or too inhomogeneous for a ‘standard’ beam from an undulator source, ∼25–50 µm (FWHM) in the vertical and 50–100 µm in the horizontal direction. Although many synchrotron facilities have microfocus beamlines for other applications, this capability for macromolecular crystallography was pioneered at ID-13 of the ESRF. The National Institute of General Medical Sciences and National Cancer Institute Collaborative Access Team (GM/CA-CAT) dual canted undulator beamlines at the APS deliver high-intensity focused beams with a minimum focal size of 20 µm × 65 µm at the sample position. To meet growing user demand for beams to study samples of 10 µm or less, a ‘mini-beam’ apparatus was developed that conditions the focused beam to either 5 µm or 10 µm (FWHM) diameter with high intensity. The mini-beam has a symmetric Gaussian shape in both the horizontal and vertical directions, and reduces the vertical divergence of the focused beam by 25%. Significant reduction in background was achieved by implementation of both forward- and back-scatter guards. A unique triple-collimator apparatus, which has been in routine use on both undulator beamlines since February 2008, allows users to rapidly interchange the focused beam and conditioned mini-beams of two sizes with a single mouse click. The device and the beam are stable over many hours of routine operation. The rapid-exchange capability has greatly facilitated sample screening and resulted in several structures that could not have been obtained with the larger focused beam. PMID:19240333

  7. Plasma spray nozzle with low overspray and collimated flow

    NASA Technical Reports Server (NTRS)

    Beason, Jr., George P. (Inventor); McKechnie, Timothy N. (Inventor); Power, Christopher A. (Inventor)

    1996-01-01

    An improved nozzle for reducing overspray in high temperature supersonic plasma spray devices comprises a body defining an internal passageway having an upstream end and a downstream end through which a selected plasma gas is directed. The nozzle passageway has a generally converging/diverging Laval shape with its upstream end converging to a throat section and its downstream end diverging from the throat section. The upstream end of the passageway is configured to accommodate a high current cathode for producing an electrical arc in the passageway to heat and ionize the gas flow to plasma form as it moves along the passageway. The downstream end of the nozzle is uniquely configured through the methodology of this invention to have a contoured bell-shape that diverges from the throat to the exit of the nozzle. Coating material in powder form is injected into the plasma flow in the region of the bell-shaped downstream end of the nozzle and the powder particles become entrained in the flow. The unique bell shape of the nozzle downstream end produces a plasma spray that is ideally expanded at the nozzle exit and thus virtually free of shock phenomena, and that is highly collimated so as to exhibit significantly reduced fanning and diffusion between the nozzle and the target. The overall result is a significant reduction in the amount of material escaping from the plasma stream in the form of overspray and a corresponding improvement in the cost of the coating operation and in the quality and integrity of the coating itself.

  8. Collimator design for a dedicated molecular breast imaging-guided biopsy system: Proof-of-concept

    SciTech Connect

    Weinmann, Amanda L.; Hruska, Carrie B.; Conners, Amy L.; O'Connor, Michael K.

    2013-01-15

    Purpose: Molecular breast imaging (MBI) is a dedicated nuclear medicine breast imaging modality that employs dual-head cadmium zinc telluride (CZT) gamma cameras to functionally detect breast cancer. MBI has been shown to detect breast cancers otherwise occult on mammography and ultrasound. Currently, a MBI-guided biopsy system does not exist to biopsy such lesions. Our objective was to consider the utility of a novel conical slant-hole (CSH) collimator for rapid (<1 min) and accurate monitoring of lesion position to serve as part of a MBI-guided biopsy system. Methods: An initial CSH collimator design was derived from the dimensions of a parallel-hole collimator optimized for MBI performed with dual-head CZT gamma cameras. The parameters of the CSH collimator included the collimator height, cone slant angle, thickness of septa and cones of the collimator, and the annular areas exposed at the base of the cones. These parameters were varied within the geometric constraints of the MBI system to create several potential CSH collimator designs. The CSH collimator designs were evaluated using Monte Carlo simulations. The model included a breast compressed to a thickness of 6 cm with a 1-cm diameter lesion located 3 cm from the collimator face. The number of particles simulated was chosen to represent the count density of a low-dose, screening MBI study acquired with the parallel-hole collimator for 10 min after a {approx}150 MBq (4 mCi) injection of Tc-99m sestamibi. The same number of particles was used for the CSH collimator simulations. In the resulting simulated images, the count sensitivity, spatial resolution, and accuracy of the lesion depth determined from the lesion profile width were evaluated. Results: The CSH collimator design with default parameters derived from the optimal parallel-hole collimator provided 1-min images with error in the lesion depth estimation of 1.1 {+-} 0.7 mm and over 21 times the lesion count sensitivity relative to 1-min images

  9. Projection space image reconstruction using strip functions to calculate pixels more "natural" for modeling the geometric response of the SPECT collimator.

    PubMed

    Hsieh, Y L; Zeng, G L; Gullberg, G T

    1998-02-01

    The spatially varying geometric response of the collimator-detector system in single photon emission computed tomography (SPECT) causes loss in resolution, shape distortions, reconstructed density nonuniformity, and quantitative inaccuracies. A projection space image reconstruction algorithm is used to correct these reconstruction artifacts. The projectors F use strip functions to calculate pixels more "natural" for modeling the two-dimensional (2-D) geometric response of the SPECT collimator transaxially to the axis of rotation. These projectors are defined by summing the intersection of an array of multiple strips rotated at equal angles to approximate the ideal system geometric response of the collimator. Two projection models were evaluated for modeling the system geometric response function. For one projector each strip is of equal weight, for the other projector a Gaussian weighting is used. Parallel beam and fan beam projections of a physical three-dimensional (3-D) Hoffman brain phantom and a Jaszczak cold rod phantom were used to evaluate the geometric response correction. Reconstructions were obtained by using the singular value decomposition (SVD) method and the iterative conjugate gradient algorithm to solve for q in the imaging equation FGq = p, where p is the projection measurement. The projector F included the new models for the geometric response, whereas, the backprojector G did not always model the geometric response in order to increase the computational speed. The final reconstruction was obtained by sampling the backprojection Gq at a discrete array of points. Reconstructions produced by the two proposed projectors showed improved resolution when compared against a unit-strip "natural" pixel model, the conventional image pixelized model with ray tracing to calculate the geometric response, and the filtered backprojection algorithm. When the reconstruction is displayed on fine grid points, the continuity and resolution of the image is preserved

  10. Thermal Photon and Residual Gas Scattering of the Electrons in the ILC RTML

    SciTech Connect

    Seletskiy, S.M.; /SLAC

    2006-08-16

    The scattering of the primary beam electrons off of thermal photons and residual gas molecules in the projected International Linear Collider (ILC) is a potential source of beam haloes which must be collimated downstream of the linac. In this report we give the analytic estimations of the individual input that each of the main scattering processes makes in the production of off-energy and large amplitude particles in the Damping Ring to Main Linac region (RTML).

  11. Influence of high-energy photons from cobalt-57 flood sources on scintillation camera uniformity images.

    PubMed

    Sokole, E B; Heckenberg, A; Bergmann, H

    1996-04-01

    Cobalt-57 flood sources are often used for system flood-field uniformity checks of scintillation camera performance. Such sources are now available in large sizes and with high activities. Uniformity images using new 57Co sources have shown artefacts which are not present in uniformity images obtained from technetium-99m flood sources of the same activity. The high-energy photons emitted by 57Co and cobalt isotope impurities appear to influence the images obtained. This was investigated in three cameras using three 57Co sources of different age. The flood-field images obtained with the 57Co sources showed non-uniformity patterns that were dependent on the age of the cobalt source, the distance of the source to the collimator, and the specific camera type. Quantification of the uniformity images reflected these findings. Energy spectra of a new 57Co source, obtained with an external 1024-channel analyser connected to the camera, showed a broad tail of high-energy photons above the 122-keV photopeak, due to Compton scatter and collimator penetration. This tail diminished with older sources and with increased source to collimator distance, indicating that in both situations fewer high-energy photons were being measured by the camera system. This tail of high-energy photons contributes to the total count rate, but because this is not obvious from the counts observed in the photopeak window, the camera can unsuspectingly be operated at too high a total countrate where pile-up effects become significant. Caution is therefore advised when using 57Co sources. Problems may be minimized by purchasing low-activity sources (certainly no more than 370 MBq), by placing the source at a distance from the collimator, e.g. 50 cm, by allowing a new source to decay, and by manufacturers producing 57Co sources without detectable radioactive contaminants.

  12. Spectra of photons and neutrons generated in a heterogeneous head of a 15 MV LINAC at differents field sizes

    SciTech Connect

    Benites-Rengifo, J. L.; Vega-Carrillo, H. R.; Velazquez-Fernandez, J. B.

    2012-10-23

    Spectra of photons and neutrons were calculated, using the Monte Carlo code MCNP-5 using the e/p/n mode. A heterogeneous model was used to define the linac head where the collimators were modeled to produce five different treatment fields at the isocenter. Photon and neutron spectra were estimated in several points along two directions from the isocenter. The total photon fluence beyond 60 cm behaves according to 1/r{sup 2} rule, while total neutron fluence, beyond 80 cm, can be described by diffusion theory using an infinite plane as a neutron source.

  13. SU-E-T-604: Dosimetric Dependence On the Collimator Angle in Prostate Volumetric Modulated Arc Therapy

    SciTech Connect

    Khan, M; Rehman, J; Khan, M; Chow, J

    2014-06-01

    Purpose: The purpose of this study is to investigate the dose-volume variations of planning target volume (PTV) and organs-at-risk (OARs) in prostate volumetric modulated arc therapy (VMAT) when using different collimator angles. It is because collimator angle awareness is essential for planner to produce an optimal prostate VMAT plan in a rational time. Methods: Single-arc VMAT plans at different collimator angles (0o, 15o, 30o, 45o, 60o, 75o and 90o) were created systematically using a Harold heterogeneous pelvis phantom. For each change of collimator angle, a new plan was re-optimized for that angle. The prescription dose was 78 Gy per 39 fractions. Conformity index (CI), homogeneity index (HI), gradient index, machine monitor unit, dose-volume histogram, the mean and maximum doses of the PTV were calculated and analyzed. On the other hand, dose-volume histogram, the mean and maximum doses of the OARs such as bladder, rectum and femoral heads for different collimator angles were determined from the plans. Results: There was no significance difference, based on the plan dose-volume evaluation criteria, found in the VMAT optimizations for all studied collimator angles. Higher CI and lower HI were found for the 45o collimator angle. In addition, the 15o collimator angle provided lower HI similar to the 45o collimator angle. The 75o and 90o collimator angle were found good for the rectum sparing, and the 75o and 30o collimator angle were found good for the right and left femur sparing, respectively. The PTV dose coverage for each plan was comparatively independent of the collimator angle. Conclusion: The dosimetric results in this study are useful to the planner to select different collimator angles to improve the PTV coverage and OAR sparing in prostate VMAT.

  14. Effect of dosimeter type for commissioning small photon beams on calculated dose distribution in stereotactic radiosurgery

    SciTech Connect

    García-Garduño, O. A. E-mail: amanda.garcia.g@gmail.com; Rodríguez-Ponce, M.; Gamboa-deBuen, I.; Rodríguez-Villafuerte, M.; Galván de la Cruz, O. O.; and others

    2014-09-15

    Purpose: To assess the impact of the detector used to commission small photon beams on the calculated dose distribution in stereotactic radiosurgery (SRS). Methods: In this study, six types of detectors were used to characterize small photon beams: three diodes [a silicon stereotactic field diode SFD, a silicon diode SRS, and a silicon diode E], an ionization chamber CC01, and two types of radiochromic film models EBT and EBT2. These detectors were used to characterize circular collimated beams that were generated by a Novalis linear accelerator. This study was conducted in two parts. First, the following dosimetric data, which are of particular interest in SRS, were compared for the different detectors: the total scatter factor (TSF), the tissue phantom ratios (TPRs), and the off-axis ratios (OARs). Second, the commissioned data sets were incorporated into the treatment planning system (TPS) to compare the calculated dose distributions and the dose volume histograms (DVHs) that were obtained using the different detectors. Results: The TSFs data measured by all of the detectors were in good agreement with each other within the respective statistical uncertainties: two exceptions, where the data were systematically below those obtained for the other detectors, were the CC01 results for all of the circular collimators and the EBT2 film results for circular collimators with diameters below 10.0 mm. The OAR results obtained for all of the detectors were in excellent agreement for all of the circular collimators. This observation was supported by the gamma-index test. The largest difference in the TPR data was found for the 4.0 mm circular collimator, followed by the 10.0 and 20.0 mm circular collimators. The results for the calculated dose distributions showed that all of the detectors passed the gamma-index test at 100% for the 3 mm/3% criteria. The aforementioned observation was true regardless of the size of the calculation grid for all of the circular collimators

  15. Transport properties of pseudospin-1 photons (Presentation Recording)

    NASA Astrophysics Data System (ADS)

    Chan, Che Ting; Fang, Anan; Zhang, Zhao-Qing; Louie, Steven G.

    2015-09-01

    Pseudospin is of central importance in governing many unusual transport properties of graphene and other artificial systems which have pseudospins of 1/2. These unconventional transport properties are manifested in phenomena such as Klein tunneling, and collimation of electron beams in one-dimensional external potentials. Here we show that in certain photonic crystals (PCs) exhibiting conical dispersions at the center of Brillouin zone, the eigenstates near the "Dirac-like point" can be described by an effective spin-orbit Hamiltonian with a pseudospin of 1. This effective Hamiltonian describes within a unified framework the wave propagations in both positive and negative refractive index media which correspond to the upper and lower conical bands respectively. Different from a Berry phase of π for the Dirac cone of pseudospin-1/2 systems, the Berry phase for the Dirac-like cone turns out to be zero from this pseudospin-1 Hamiltonian. In addition, we found that a change of length scale of the PC can shift the Dirac-like cone rigidly up or down in frequency with its group velocity unchanged, hence mimicking a gate voltage in graphene and allowing for a simple mechanism to control the flow of pseudospin-1 photons. As a photonic analogue of electron potential, the length-scale induced Dirac-like point shift is effectively a photonic potential within the effective pseudospin-1 Hamiltonian description. At the interface of two different potentials, the 3-component spinor gives rise to distinct boundary conditions which do not require each component of the wave function to be continuous, leading to new wave transport behaviors as shown in Klein tunneling and supercollimation. For examples, the Klein tunneling of pseudospin-1 photons is much less anisotropic with reference to the incident angle than that of pseudospin-1/2 electrons, and collimation can be more robust with pseudospin-1 than pseudospin-1/2. The special wave transport properties of pseudospin-1 photons

  16. RR photons

    NASA Astrophysics Data System (ADS)

    Cámara, Pablo G.; Ibáñez, Luis E.; Marchesano, Fernando

    2011-09-01

    Type II string compactifications to 4d generically contain massless Ramond-Ramond U(1) gauge symmetries. However there is no massless matter charged under these U(1)'s, which makes a priori difficult to measure any physical consequences of their existence. There is however a window of opportunity if these RR U(1)'s mix with the hypercharge U(1) Y (hence with the photon). In this paper we study in detail different avenues by which U(1) RR bosons may mix with D-brane U(1)'s. We concentrate on Type IIA orientifolds and their M-theory lift, and provide geometric criteria for the existence of such mixing, which may occur either via standard kinetic mixing or via the mass terms induced by Stückelberg couplings. The latter case is particularly interesting, and appears whenever D-branes wrap torsional p-cycles in the compactification manifold. We also show that in the presence of torsional cycles discrete gauge symmetries and Aharanov-Bohm strings and particles appear in the 4d effective action, and that type IIA Stückelberg couplings can be understood in terms of torsional (co)homology in M-theory. We provide examples of Type IIA Calabi-Yau orientifolds in which the required torsional cycles exist and kinetic mixing induced by mass mixing is present. We discuss some phenomenological consequences of our findings. In particular, we find that mass mixing may induce corrections relevant for hypercharge gauge coupling unification in F-theory SU(5) GUT's.

  17. Photon-Photon Collisions -- Past and Future

    SciTech Connect

    Brodsky, Stanley J.; /SLAC

    2005-12-02

    I give a brief review of the history of photon-photon physics and a survey of its potential at future electron-positron colliders. Exclusive hadron production processes in photon-photon and electron-photon collisions provide important tests of QCD at the amplitude level, particularly as measures of hadron distribution amplitudes. There are also important high energy {gamma}{gamma} and e{gamma} tests of quantum chromodynamics, including the production of jets in photon-photon collisions, deeply virtual Compton scattering on a photon target, and leading-twist single-spin asymmetries for a photon polarized normal to a production plane. Since photons couple directly to all fundamental fields carrying the electromagnetic current including leptons, quarks, W's and supersymmetric particles, high energy {gamma}{gamma} collisions will provide a comprehensive laboratory for Higgs production and exploring virtually every aspect of the Standard Model and its extensions. High energy back-scattered laser beams will thus greatly extend the range of physics of the International Linear Collider.

  18. Relative performance of a TGS for the assay of drummed waste as function of collimator opening

    SciTech Connect

    Kane, S.C.; Croft, S.; McClay, P.; Venkataraman, R.; Villani, M.F.

    2007-07-01

    Improving the safety, accuracy and overall cost effectiveness of the processes and methods used to characterize and handle radioactive waste is an on-going mission for the nuclear industry. An important contributor to this goal is the development of superior non-destructive assay instruments. The Tomographic Gamma Scanner (TGS) is a case in point. The TGS applies low spatial resolution experimental computed tomography (CT) linear attenuation coefficient maps with three-dimensional high-energy resolution single photon emission reconstructions. The results are presented as quantitative matrix attenuation corrected images and assay values for gamma-emitting radionuclides. Depending on a number of operational factors, this extends the diversity of waste forms that can be assayed, to a given accuracy, to items containing more heterogeneous matrix distributions and less uniform emission activity distributions. Recent advances have significantly extended the capability to a broader range of matrix density and to a wider dynamic range of surface dose rate. Automated systems sense the operational conditions, including the container type, and configure themselves accordingly. The TGS also provides a flexible data acquisition platform and can be used to perform far-field style measurements, classical segmented gamma scanner measurements, or to implement hybrid methods, such as reconstructions that use a priori knowledge to constrain the image reconstruction or the underlying energy dependence of the attenuation. A single, yet flexible, general purpose instrument of this kind adds several tiers of strategic and tactical value to facilities challenged by a diverse and difficult to assay waste streams. The TGS is still in the early phase of industrial uptake. There are only a small number of general purpose TGS systems operating worldwide, most being configured to automatically select between a few configurations appropriate for routine operations. For special investigations

  19. Stripe pattern in the intensity profile of collimated soft x-ray beams caused by surface corrugation of the refocusing mirrors

    NASA Astrophysics Data System (ADS)

    Schmitz, D.; Siewert, F.; Zeschke, T.

    2015-02-01

    The effect of progress in surface finishing of optical components on the collimated-beam properties of soft x-ray beamlines at synchrotron radiation facilities is demonstrated: a stripe pattern, experimentally observed in the 2D intensity profile of beamlines with optical components manufactured 10-15 years ago, would be strongly attenuated if the existing refocusing mirror was replaced by an ultra-precise mirror manufactured with state-of-the-art of today surface finishing techniques. The observed stripe pattern is not caused by diffraction because its period length did not change with photon energy. Instead it can be explained with geometrical optics and is due to the height profile of the refocusing mirror which has been independently measured with a long trace profiler and used as an input in our raytracing simulations.

  20. Nuclear photonics

    SciTech Connect

    Habs, D.; Guenther, M. M.; Jentschel, M.; Thirolf, P. G.

    2012-07-09

    With the planned new {gamma}-beam facilities like MEGa-ray at LLNL (USA) or ELI-NP at Bucharest (Romania) with 10{sup 13}{gamma}/s and a band width of {Delta}E{gamma}/E{gamma} Almost-Equal-To 10{sup -3}, a new era of {gamma} beams with energies up to 20MeV comes into operation, compared to the present world-leading HI{gamma}S facility at Duke University (USA) with 10{sup 8}{gamma}/s and {Delta}E{gamma}/E{gamma} Almost-Equal-To 3 Dot-Operator 10{sup -2}. In the long run even a seeded quantum FEL for {gamma} beams may become possible, with much higher brilliance and spectral flux. At the same time new exciting possibilities open up for focused {gamma} beams. Here we describe a new experiment at the {gamma} beam of the ILL reactor (Grenoble, France), where we observed for the first time that the index of refraction for {gamma} beams is determined by virtual pair creation. Using a combination of refractive and reflective optics, efficient monochromators for {gamma} beams are being developed. Thus, we have to optimize the total system: the {gamma}-beam facility, the {gamma}-beam optics and {gamma} detectors. We can trade {gamma} intensity for band width, going down to {Delta}E{gamma}/E{gamma} Almost-Equal-To 10{sup -6} and address individual nuclear levels. The term 'nuclear photonics' stresses the importance of nuclear applications. We can address with {gamma}-beams individual nuclear isotopes and not just elements like with X-ray beams. Compared to X rays, {gamma} beams can penetrate much deeper into big samples like radioactive waste barrels, motors or batteries. We can perform tomography and microscopy studies by focusing down to {mu}m resolution using Nuclear Resonance Fluorescence (NRF) for detection with eV resolution and high spatial resolution at the same time. We discuss the dominating M1 and E1 excitations like the scissors mode, two-phonon quadrupole octupole excitations, pygmy dipole excitations or giant dipole excitations under the new facet of

  1. Nuclear photonics

    NASA Astrophysics Data System (ADS)

    Habs, D.; Günther, M. M.; Jentschel, M.; Thirolf, P. G.

    2012-07-01

    With the planned new γ-beam facilities like MEGa-ray at LLNL (USA) or ELI-NP at Bucharest (Romania) with 1013 γ/s and a band width of ΔEγ/Eγ≈10-3, a new era of γ beams with energies up to 20MeV comes into operation, compared to the present world-leading HIγS facility at Duke University (USA) with 108 γ/s and ΔEγ/Eγ≈3ṡ10-2. In the long run even a seeded quantum FEL for γ beams may become possible, with much higher brilliance and spectral flux. At the same time new exciting possibilities open up for focused γ beams. Here we describe a new experiment at the γ beam of the ILL reactor (Grenoble, France), where we observed for the first time that the index of refraction for γ beams is determined by virtual pair creation. Using a combination of refractive and reflective optics, efficient monochromators for γ beams are being developed. Thus, we have to optimize the total system: the γ-beam facility, the γ-beam optics and γ detectors. We can trade γ intensity for band width, going down to ΔEγ/Eγ≈10-6 and address individual nuclear levels. The term "nuclear photonics" stresses the importance of nuclear applications. We can address with γ-beams individual nuclear isotopes and not just elements like with X-ray beams. Compared to X rays, γ beams can penetrate much deeper into big samples like radioactive waste barrels, motors or batteries. We can perform tomography and microscopy studies by focusing down to μm resolution using Nuclear Resonance Fluorescence (NRF) for detection with eV resolution and high spatial resolution at the same time. We discuss the dominating M1 and E1 excitations like the scissors mode, two-phonon quadrupole octupole excitations, pygmy dipole excitations or giant dipole excitations under the new facet of applications. We find many new applications in biomedicine, green energy, radioactive waste management or homeland security. Also more brilliant secondary beams of neutrons and positrons can be produced.

  2. On the selection of gantry and collimator angles for isocenter localization using Winston-Lutz tests.

    PubMed

    Du, Weiliang; Johnson, Jennifer L; Jiang, Wei; Kudchadker, Rajat J

    2016-01-01

    In Winston-Lutz (WL) tests, the isocenter of a linear accelerator (linac) is determined as the intersection of radiation central axes (CAX) from multiple gantry, collimator, and couch angles. It is well known that the CAX can wobble due to mechanical imperfections of the linac. Previous studies suggested that the wobble varies with gantry and collimator angles. Therefore, the isocenter determined in the WL tests has a profound dependence on the gantry and collimator angles at which CAX are sampled. In this study, we evaluated the systematic and random errors in the iso-centers determined with different CAX sampling schemes. Digital WL tests were performed on six linacs. For each WL test, 63 CAX were sampled at nine gantry angles and seven collimator angles. Subsets of these data were used to simulate the effects of various CAX sampling schemes. An isocenter was calculated from each subset of CAX and compared against the reference isocenter, which was calculated from 48 opposing CAX. The differences between the calculated isocenters and the reference isocenters ranged from 0 to 0.8 mm. The differences diminished to less than 0.2 mm when 24 or more CAX were sampled. Isocenters determined with collimator 0° were vertically lower than those determined with collimator 90° and 270°. Isocenter localization errors in the longitudinal direction (along the axis of gantry rotation) showed a strong dependence on the collimator angle selected. The errors in all directions were significantly reduced when opposing collimator angles and opposing gantry angles were employed. The isocenter localization errors were less than 0.2 mm with the common CAX sampling scheme, which used four cardinal gantry angles and two opposing collimator angles. Reproducibility stud-ies on one linac showed that the mean and maximum variations of CAX during the WL tests were 0.053 mm and 0.30 mm, respectively. The maximal variation in the resulting isocenters was 0.068 mm if 48 CAX were used, or 0

  3. Modeling and design of micromachined optical Söller collimators for lensless CCD-based fluorometry.

    PubMed

    Balsam, Joshua; Ossandon, Miguel; Bruck, Hugh Alan; Rasooly, Avraham

    2012-11-01

    To address the needs of medical diagnostics in resource-poor settings, it is necessary to develop low cost, simple and portable Point of Care detectors for integrated medical diagnostics. Previously, we have described a simple lensless fluorometer with sensitivity in the range of current ELISA plate readers. The key to the lensfree fluorometer is the uniform spatial distribution of light, which we achieved using a simple optical collimator based on a "stack of pinholes" (a stack of black PMMA plates with arrays of pinholes machined via laser) enabling the light to be collimated from the LED light source through the necessary wavelength filters and the assay's microfluidics directly onto the CCD without a lens. In this paper, we describe the optical principle for designing these Söller collimators for lensfree CCD-based fluorometry. The illuminating surface was modeled as a collection of differential areas emitting uniformly and spherically, and the intensity contribution of each emitting area was summed over the detector surface. To compute the final light intensity distribution from such a differential model we derived an integral equation to sum the individual intensity contributions from the two-dimensional emitting surface. The equation is for a single-hole collimator. Light intensity measurements were taken by placing a collimator with a particular aspect ratio (the ratio of hole length to diameter (L/d)) over the CCD image sensor and capturing an image. The resulting image is the 2D light intensity profile generated by the collimator. As the aspect ratio is increased the slope of the light intensity profile increases, corresponding to an increased degree of collimation. To test the model, the measured maximum and mean light intensities were compared with the theoretical predictions generated from the model. There was an agreement between the variation of the mean (R(2) = 0.990) and maximum (R(2) = 0.938) values of light intensities with aspect ratios based

  4. Strict X-ray beam collimation for facial bones examination can increase lens exposure

    PubMed Central

    Powys, R; Robinson, J; Kench, P L; Ryan, J; Brennan, P C

    2012-01-01

    Objectives It is well accepted that collimation is a cost-effective dose-reducing tool for X-ray examinations. This phantom-based study investigated the impact of X-ray beam collimation on radiation dose to the lenses of the eyes and thyroid along with the effect on image quality in facial bone radiography. Methods A three-view series (occipitomental, occipitomental 30 and lateral) was investigated, and radiation doses to the lenses and thyroid were measured using an Unfors dosemeter. Images were assessed by six experienced observers using a visual grading analysis and a total of 5400 observations were made. Results Strict collimation significantly (p<0.0001) reduced the radiation dose to the lenses of the eyes and thyroid when using a fixed projection-specific exposure. With a variable exposure technique (fixed exit dose, to simulate the behaviour of an automatic exposure control), while strict collimation was again shown to reduce thyroid dose, higher lens doses were demonstrated when compared with larger fields of exposure. Image quality was found to significantly improve using strict collimation, with observer preference being demonstrated using visual grading characteristic curves. Conclusion The complexities of optimising radiographic techniques have been shown and the data presented emphasise the importance of examining dose-reducing strategies in a comprehensive way. PMID:22374279

  5. Monolithic fiber end cap collimator for high-power free-space fiber-fiber coupling.

    PubMed

    Zhou, Xuanfeng; Chen, Zilun; Wang, Zefeng; Hou, Jing

    2016-05-20

    In this paper, we present the design, construction, and testing of a monolithic fiber end cap collimator for high-power free-space fiber-fiber coupling applications. The collimator is based on a large-sized fiber end cap and a spherical lens design on the output facet. Values of the spot size and working distance are theoretically analyzed based on Gaussian approximation and ABCD transmission matrix. The free-space fiber-fiber coupling process is also analyzed for different lens curvature radii and coupling distances. In the experiment, a collimated laser beam is obtained with Rayleigh length of about 400 mm. A high-power laser with 1.1 kW output is tested on the end cap collimator, which only heats up by 7°C at the output facet without active cooling. Free-space fiber-fiber coupling between two 20/400 μm fibers is achieved based on these collimators, with measured coupling loss lower than 0.3 dB. PMID:27411125

  6. Evaluation of Beam Loss and Energy Depositions for a Possible Phase II Design for LHC Collimation

    SciTech Connect

    Lari, L.; Assmann, R.; Bracco, C.; Brugger, M.; Cerutti, F.; Doyle, E.; Ferrari, A.; Keller, L.; Lundgren, S.; Markiewicz, Thomas W.; Mauri, M.; Redaelli, S.; Sarchiapone, L.; Smith, J.; Vlachoudis, V.; Weiler, T.; /CERN

    2011-11-07

    The LHC beams are designed to have high stability and to be stored for many hours. The nominal beam intensity lifetime is expected to be of the order of 20h. The Phase II collimation system has to be able to handle particle losses in stable physics conditions at 7 TeV in order to avoid beam aborts and to allow correction of parameters and restoration to nominal conditions. Monte Carlo simulations are needed in order to evaluate the behavior of metallic high-Z collimators during operation scenarios using a realistic distribution of losses, which is a mix of the three limiting halo cases. Moreover, the consequences in the IR7 insertion of the worst (case) abnormal beam loss are evaluated. The case refers to a spontaneous trigger of the horizontal extraction kicker at top energy, when Phase II collimators are used. These studies are an important input for engineering design of the collimation Phase II system and for the evaluation of their effect on adjacent components. The goal is to build collimators that can survive the expected conditions during LHC stable physics runs, in order to avoid quenches of the SC magnets and to protect other LHC equipments.

  7. Measurements and analysis of a high-brightness electron beam collimated in a magnetic bunch compressor

    NASA Astrophysics Data System (ADS)

    Zhou, F.; Bane, K.; Ding, Y.; Huang, Z.; Loos, H.; Raubenheimer, T.

    2015-05-01

    A collimator located in a magnetic bunch compressor of a linear accelerator driven x-ray free electron laser has many potential applications, such as the removal of horns in the current distribution, the generation of ultrashort beams, and as a diagnostic of the beam slice emittance. Collective effects, however, are a major concern in applying the technique. Systematic measurements of emittance and analysis were performed using a collimator in the first bunch compressor of the Linac Coherent Light Source (LCLS). In the nominal, undercompressed configuration using the collimator we find that the y emittance (nonbending plane) is not increased, and the x emittance (in the bending plane) is increased by about 25%, in comparison to the injector emittance. From the analysis we conclude that the parasitic effects associated with this method are dominated by coherent synchrotron radiation (CSR), which causes a "systematic error" for measuring slice emittance at the bending plane using the collimation method. In general, we find good agreement between the measurements and simulations including CSR. However, for overcompressed beams at smaller collimator gaps, an extra emittance increase is found that does not agree with 1D simulations and is not understood.

  8. Medical linear accelerator mounted mini-beam collimator: design, fabrication and dosimetric characterization.

    PubMed

    Cranmer-Sargison, G; Crewson, C; Davis, W M; Sidhu, N P; Kundapur, V

    2015-09-01

    The goal of this work was to design, build and experimentally characterize a linear accelerator mounted mini-beam collimator for use at a nominal 6 MV beam energy. Monte Carlo simulation was used in the design and dosimetric characterization of a compact mini-beam collimator assembly mounted to a medical linear accelerator. After fabrication, experimental mini-beam dose profiles and central axis relative output were measured and the results used to validate the simulation data. The simulation data was then used to establish traceability back to an established dosimetric code of practice. The Monte Carlo simulation work revealed that changes in collimator blade width have a greater influence on the valley-to-peak dose ratio than do changes in blade height. There was good agreement between the modeled and measured profile data, with the exception of small differences on either side of the central peak dose. These differences were found to be systematic across all depths and result from limitations associated with the collimator fabrication. Experimental mini-beam relative output and simulation data agreed to better than ± 2.0%, which is well within the level of uncertainty required for dosimetric traceability of non-standard field geometries. A mini-beam collimator has now been designed, built and experimentally characterized for use with a commercial linear accelerator operated at a nominal 6 MV beam energy.

  9. Collimator angle influence on dose distribution optimization for vertebral metastases using volumetric modulated arc therapy

    SciTech Connect

    Mancosu, Pietro; Cozzi, Luca; Fogliata, Antonella; Lattuada, Paola; Reggiori, Giacomo; Cantone, Marie Claire; Navarria, Pierina; Scorsetti, Marta

    2010-08-15

    Purpose: The cylindrical symmetry of vertebrae favors the use of volumetric modulated arc therapy in generating a dose ''hole'' on the center of the vertebrae limiting the dose to the spinal cord. The authors have evaluated if collimator angle is a significant parameter for dose distribution optimization in vertebral metastases. Methods: Three patients with one-three vertebrae involved were considered. Twenty-one differently optimized plans (nine single-arc and 12 double-arc plans) were performed, testing various collimator angle positions. Clinical target volume was defined as the whole vertebrae, excluding the spinal cord canal. The planning target volume (PTV) was defined as CTV+5 mm. Dose prescription was 5x4 Gy{sup 2} with normalization to PTV mean dose. The dose at 1 cm{sup 3} of spinal cord was limited to 11.5Gy. Results: The best plans in terms of target coverage and spinal cord sparing were achieved by two arcs and Arc1-80 deg. and Arc2-280 deg. collimator angles for all the cases considered (i.e., leaf travel parallel to the spinal cord primary orientation). If one arc is used, only 80 deg. reached the objectives. Conclusions: This study demonstrated the role of collimation rotation for the vertebrae metastasis irradiation, with the leaf travel parallel to the spinal cord primary orientation to be better than other solutions. Thus, optimal choice of collimator angle increases the optimization freedom to shape a desired dose distribution.

  10. Medical linear accelerator mounted mini-beam collimator: design, fabrication and dosimetric characterization.

    PubMed

    Cranmer-Sargison, G; Crewson, C; Davis, W M; Sidhu, N P; Kundapur, V

    2015-09-01

    The goal of this work was to design, build and experimentally characterize a linear accelerator mounted mini-beam collimator for use at a nominal 6 MV beam energy. Monte Carlo simulation was used in the design and dosimetric characterization of a compact mini-beam collimator assembly mounted to a medical linear accelerator. After fabrication, experimental mini-beam dose profiles and central axis relative output were measured and the results used to validate the simulation data. The simulation data was then used to establish traceability back to an established dosimetric code of practice. The Monte Carlo simulation work revealed that changes in collimator blade width have a greater influence on the valley-to-peak dose ratio than do changes in blade height. There was good agreement between the modeled and measured profile data, with the exception of small differences on either side of the central peak dose. These differences were found to be systematic across all depths and result from limitations associated with the collimator fabrication. Experimental mini-beam relative output and simulation data agreed to better than ± 2.0%, which is well within the level of uncertainty required for dosimetric traceability of non-standard field geometries. A mini-beam collimator has now been designed, built and experimentally characterized for use with a commercial linear accelerator operated at a nominal 6 MV beam energy. PMID:26305166

  11. Strengthening weak-value amplification with recycled photons

    NASA Astrophysics Data System (ADS)

    Dressel, Justin; Lyons, Kevin; Jordan, Andrew N.; Graham, Trent M.; Kwiat, Paul G.

    2013-08-01

    We consider the use of cyclic weak measurements to improve the sensitivity of weak-value amplification precision measurement schemes. Previous weak-value experiments have used only a small fraction of events, while discarding the rest through the process of “postselection.” We extend this idea by considering recycling of events which are typically unused in a weak measurement. Here we treat a sequence of polarized laser pulses effectively trapped inside an interferometer using a Pockels cell and polarization optics. In principle, all photons can be postselected, which will improve the measurement sensitivity. We first provide a qualitative argument for the expected improvements from recycling photons, followed by the exact result for the recycling of collimated beam pulses, and numerical calculations for diverging beams. We show that beam degradation effects can be mitigated via profile flipping or Zeno reshaping. The main advantage of such a recycling scheme is an effective power increase, while maintaining an amplified deflection.

  12. Controllable photon source

    NASA Astrophysics Data System (ADS)

    Oszetzky, Dániel; Nagy, Attila; Czitrovszky, Aladár

    2006-10-01

    We have developed our pervious experimental setup using correlated photon pairs (to the calibration of photo detectors) to realize a controllable photon source. For the generation of such photon pairs we use the non-linear process of parametric down conversion. When a photon of the pump beam is incident to a nonlinear crystal with phase matching condition, a pair of photons (signal and idler) is created at the same time with certain probability. We detect the photons in the signal beam with a single photon counting module (SPCM), while delaying those in the idler beam. Recently we have developed a fast electronic unit to control an optical shutter (a Pockels cell) placed to the optical output of the idler beam. When we detect a signal photon with the controlling electronic unit we are also able to open or close the fast optical shutter. Thus we can control which idler photons can propagate through the Pockels cell. So with this photon source we are able to program the number of photons in a certain time window. This controllable photon source that is able to generate a known number of photons with specified wavelength, direction, and polarization could be useful for applications in high-accuracy optical characterisation of photometric devices at the ultra-low intensities. This light source can also serve as a standard in testing of optical image intensifiers, night vision devices, and in the accurate measurement of spectral distribution of transmission and absorption in optical materials.

  13. A scheme design of collimator for gantry in proton therapy facility

    NASA Astrophysics Data System (ADS)

    Liu, Ming; Yin, Chongxian; Chu, Kecheng; Zhao, Liying; Shu, Hang; Dai, Xiaolei; Zhang, Manzhou; Miao, Chunhui

    2015-08-01

    In proton radiotherapy, a round dose profile at the isocenter is beneficial to the treatment planning in the active scanning mode. Based on the theory of the round-beam method, a collimator was designed to be placed at the entrance of the gantry. A simulation of Monte Carlo method was done to validate the design scheme. With the constraining function of the collimator, the dose profile at the isocenter would maintain a round shape despite of an initial asymmetric phase space in extraction of synchrotron. Furthermore, the shape of dose profile is irrelevant to the rotating angle of the gantry. The mechanical error caused by the axis deviation of the gantry could also be partly eliminated with the collimator.

  14. Arrayed waveguide collimators for integrating free-space optics on polymeric waveguide devices.

    PubMed

    Shin, Jin-Soo; Lee, Chang-Hee; Shin, Sang-Yung; Huang, Guang-Hao; Chu, Woo-Sung; Oh, Min-Cheol; Noh, Young-Ouk; Lee, Hyung-Jong

    2014-10-01

    Array-type optical devices are important for wavelength-division multiplexing optical communication system to achieve small footprint, mass production, and reliability. For fabricating transmitter module in an array configuration, it is difficult to achieve a passive alignment of isolator, collimating lens, and laser diode. To facilitate array isolator integration, a waveguide collimator is proposed in this work by using a low-contrast, large-core polymer waveguide. The diffraction of a guided mode propagating through a free-space region is suppressed by enlarging the guided mode. The fiber coupling loss due to the enlarged mode was overcome by incorporating an adiabatic taper structure. The excess loss of waveguide collimator including the loss through a 400-μm free-propagation region was less than 1.0 dB. PMID:25321959

  15. Thomson scattering measurement of a collimated plasma jet generated by a high-power laser system

    NASA Astrophysics Data System (ADS)

    Ishikawa, T.; Sakawa, Y.; Morita, T.; Yamaura, Y.; Kuramitsu, Y.; Moritaka, T.; Sano, T.; Shimoda, R.; Tomita, K.; Uchino, K.; Matsukiyo, S.; Mizuta, A.; Ohnishi, N.; Crowston, R.; Woolsey, N.; Doyle, H.; Gregori, G.; Koenig, M.; Michaut, C.; Pelka, A.; Yuan, D.; Li, Y.; Zhang, K.; Zhong, J.; Wang, F.; Takabe, H.

    2016-03-01

    One of the important and interesting problems in astrophysics and plasma physics is collimation of plasma jets. The collimation mechanism, which causes a plasma flow to propagate a long distance, has not been understood in detail. We have been investigating a model experiment to simulate astrophysical plasma jets with an external magnetic field [Nishio et al., EPJ. Web of Conferences 59, 15005 (2013)]. The experiment was performed by using Gekko XII HIPER laser system at Institute of Laser Engineering, Osaka University. We shot CH plane targets (3 mm × 3 mm × 10 μm) and observed rear-side plasma flows. A collimated plasma flow or plasma jet was generated by separating focal spots of laser beams. In this report, we measured plasma jet structure without an external magnetic field with shadowgraphy, and simultaneously measured the local parameters of the plasma jet, i.e., electron density, electron and ion temperatures, charge state, and drift velocity, with collective Thomson scattering.

  16. Beam collimation and machine detector interface at the International Linear Collider

    SciTech Connect

    Mokhov, N.V.; Drozhdin, A.I.; Kostin, M.A.; /Fermilab

    2005-05-01

    Synchrotron radiation, beam-gas scattering and beam halo interactions with collimators and other components in the ILC beam delivery system (BDS) would create fluxes of muons and other secondaries which could exceed the tolerable levels at a detector by a few orders of magnitude. It is shown that with a multi-stage collimation system, magnetized iron spoilers which fill the tunnel and a set of masks in the detector, one can hopefully meet the design goals. Results of modeling with the STRUCT and MARS15 codes of beam loss and energy deposition effects are presented in this paper. We focus on the collimation system and mask performance optimization, short- and long-term survivability of the critical components (spoilers, absorbers and magnets), dynamic heat loads and radiation levels in magnets and other components, and machine-related backgrounds in collider detectors.

  17. Long-range visible light communication system based on LED collimating lens

    NASA Astrophysics Data System (ADS)

    Chen, Yingcong; Wen, Shangsheng; Wu, Yuxiang; Ren, Yuanyuan; Guan, Weipeng; Zhou, Yunlin

    2016-10-01

    An advanced visible light communication (VLC) system is proposed for long-range VLC, such as marine communication. The design of the system is conducted into two parts. Firstly, we design and optimize a collimating lens for the optical antenna by using Taguchi method. The lighting effects and optical power of the receiving end in different distances are simulated by TracePro software. Then, the long-range VLC channel is reconstructed by integrating the influence of the atmospheric attenuation and frequency response. The performance of the OOK coding VLC system is tested by Matlab software. The results show that: the emitting angle of the optimized collimating lens is 1.7°. By using 1 W LED and collimating lens as an optical antenna, the system can achieve a data rate of 210 Mbit/s at a bit error rate of 10-3 in 90 m.

  18. Stellar physics. Observing the onset of outflow collimation in a massive protostar.

    PubMed

    Carrasco-González, C; Torrelles, J M; Cantó, J; Curiel, S; Surcis, G; Vlemmings, W H T; van Langevelde, H J; Goddi, C; Anglada, G; Kim, S-W; Kim, J-S; Gómez, J F

    2015-04-01

    The current paradigm of star formation through accretion disks, and magnetohydrodynamically driven gas ejections, predicts the development of collimated outflows, rather than expansion without any preferential direction. We present radio continuum observations of the massive protostar W75N(B)-VLA 2, showing that it is a thermal, collimated ionized wind and that it has evolved in 18 years from a compact source into an elongated one. This is consistent with the evolution of the associated expanding water-vapor maser shell, which changed from a nearly circular morphology, tracing an almost isotropic outflow, to an elliptical one outlining collimated motions. We model this behavior in terms of an episodic, short-lived, originally isotropic ionized wind whose morphology evolves as it moves within a toroidal density stratification. PMID:25838383

  19. Large area x-ray collimator-the zone plate approach.

    PubMed

    Menz, Benedikt; Braig, Christoph; Bräuninger, Heinrich; Burwitz, Vadim; Hartner, Gisela; Predehl, Peter

    2015-09-10

    One question of particular interest in the measurement of x-ray imaging optics for space telescopes concerns the characteristics of the point spread function (PSF) in orbit and the focal length for an infinite source distance. In order to measure such a PSF, a parallel x-ray beam with a diameter of several centimeters to meters is required. For this purpose a large area transmission x-ray zone plate (ZP) for collimating x-ray beams has been designed, built, and tested. Furthermore we present a setup to determine large-scale aberrations of the collimated beam. From x-ray measurements we obtain an upper limit for the angular resolution of ±0.2 arc sec and a first-order diffraction efficiency of ≈13%. These results show that it is possible to use a ZP as a collimator for the PANTER x-ray test facility. PMID:26368954

  20. Stellar physics. Observing the onset of outflow collimation in a massive protostar.

    PubMed

    Carrasco-González, C; Torrelles, J M; Cantó, J; Curiel, S; Surcis, G; Vlemmings, W H T; van Langevelde, H J; Goddi, C; Anglada, G; Kim, S-W; Kim, J-S; Gómez, J F

    2015-04-01

    The current paradigm of star formation through accretion disks, and magnetohydrodynamically driven gas ejections, predicts the development of collimated outflows, rather than expansion without any preferential direction. We present radio continuum observations of the massive protostar W75N(B)-VLA 2, showing that it is a thermal, collimated ionized wind and that it has evolved in 18 years from a compact source into an elongated one. This is consistent with the evolution of the associated expanding water-vapor maser shell, which changed from a nearly circular morphology, tracing an almost isotropic outflow, to an elliptical one outlining collimated motions. We model this behavior in terms of an episodic, short-lived, originally isotropic ionized wind whose morphology evolves as it moves within a toroidal density stratification.